United States
Environmental Protection
Agency
Office of Solid Waste and
Emergency Response
Washington DC 20460
EPAy540/P-91/006
January 1991
Compendium of ERT
Soil Sampling and
Surface Geophsics
Procedures

-------

-------
                                                 EPA/540/P-91/006
                                            OSWER Directive 9360.4-02
                                                     January 1991
COMPENDIUM OF ERT SOIL SAMPLING AND
    SURFACE GEOPHYSICS PROCEDURES
               Sampling Equipment Decontamination

               Soil Sampling

               Soil Gas Sampling

               General Surface Geophysics
                       Interim Final
                  Environmental Response Team
                  Emergency Response Division
              Office of Emergency and Remedial Response
                U.S. Environmental Protection Agency
                    Washington, DC 20460
                                             Printed on Recycled Paper

-------
                                             Notice
This document has been reviewed in accordance with U.S. Environmental Protection Agency policy and approved
for publication.  Mention of trade names or commercial  products does  not constitute  endorsement or
recommendation for use.

The policies and procedures established in this document are intended solely for the guidance of government
personnel for use in the Superfund Removal Program. They are not intended, and cannot be relied upon, to
create any rights, substantive or procedural, enforceable by any party in litigation with the United States.  The
Agency reserves the right to act at variance with these policies and procedures and to change them at any time
without public notice.

Depending on circumstances and needs, it may not be possible or appropriate to follow these procedures exactly
in all situations  due to site conditions, equipment limitations, and limitations of the standard procedures.
Whenever these procedures cannot be followed as written, they may be used as general guidance with any and
all modifications fully documented in either QA Plans, Sampling Plans, or final reports of results.

Each Standard Operating Procedure in this compendium contains  a discussion on quality assurance/quality
control (QA/QC).  For more information on QA/QC objectives and requirements,  refer to the Quality
Assurance/Quality Control Guidance for Removal Activities, OSWER directive 9360.4-01, EPA/540/G-90/004.

Questions, comments, and recommendations are welcomed regarding the Compendium of ERT Soil Sampling
and Surface Geophysics Procedures. Send remarks to:

                                      Mr. William A. Coakley
                                 Removal Program QA Coordinator
                                          U.S. EPA - ERT
                                 Raritan Depot - Building 18, MS-101
                                      2890 Woodbridge Avenue
                                        Edison, NJ 08837-3679

For additional copies of the Compendium of ERT Soil Sampling and Surface Geophysics Procedures, please
contact:

                            National Technical Information Service (NTIS)
                                    U.S. Department of Commerce
                                        5285 Port Royal Road
                                        Springfield, VA 22161
                                           (703) 487-4600

-------
                                       Table of Contents
 1.0     SAMPLING EQUIPMENT DECONTAMINATION: SOP #2006

        1.1     Scope and Application                                                              1
        1.2     Method Summary                                                                  1
        1.3     Sample Preservation, Containers, Handling, and Storage                                1
        1.4     Interferences and Potential Problems                                                 1
        1.5     Equipment/Apparatus                                                              1
        1.6     Reagents                                                                          2
        1.7     Procedures                                                                        2

               1.7.1    Decontamination Methods                                                   2
               1.7.2    Field Sampling Equipment Cleaning Procedures                                3

        1.8     Calculations                                                                       3
        1.9     Quality Assurance/Quality Control                                                   3
        1.10    Data Validation                                                                    4
        1.11    Health and Safety                                                                  4


2.0     SOIL SAMPLING: SOP #2012

        2.1     Scope and Application                                                              5
        2.2     Method Summary                                                                  5
        2.3     Sample Preservation,  Containers, Handling, and Storage                                5
        2.4     Interferences and Potential Problems                                                 5
        2.5     Equipment/Apparatus                                                              5
        2.6     Reagents                                                                          5
        2.7     Procedures                                                                        6

               2.7.1   Preparation                                                                6
               2.7.2   Sample Collection                                                          6

        2.8     Calculations                                                                       9
        2.9     Quality Assurance/Quality Control                                                   9
        2.10    Data Validation                                                                    9
        2.11    Health and Safety                                                                  9
3.0     SOIL GAS SAMPLING:  SOP #2149

       3.1     Scope and Application                                                            n
       3.2     Method Summary                                                                U
       3.3     Sample Preservation, Containers, Handling, and Storage                               11

               3.3.1    Tedlar Bag                                                               11
               3.3.2    TenaxTube                                                              11
               3.3.3    SUMMA Canister                                                        11
                                                111

-------
Section
        3.4     Interferences and Potential Problem

               3.41    HNU Measurements 12
               3.4.2    Factors Affecting Organic Concentrations in Soil Gas
               3.43    Soil Probe Clogging
               3.4.4    Underground Utilities

        3.5     Equipment/Apparatus

               3.5JL    Slam Bar Method
               3.5.2    Power Hammer Method

        3.6     Reagents
        3.7     Procedures

               3.71    Soil Gas Well Installation
               3.7.2    Screening with Field Instruments
               3.73    Tedlar Bag Sampling
               3.7.4    Tenax Tube Sampling
               3.7.5    SUMMA Canister Sampling

        3.8     Calculations

               3.8JL    Field Screening Instruments
               3.8.2    Photovac GC Analysis

        3.9     Quality Assurance/Quality Control

               3.91    Field Instrument Calibration
               3.9.2    Gilian Model HFS113A Air Sampling Pump Calibration
               3.9.3    Sample Probe Contamination
               3.9.4    Sample Train Contamination
               3.9.5    Field Blank
               3.9.6    Trip Standard
               3.9.7    Tedlar Bag Check
               3.9.8    SUMMA Canister Check
               3.9.9    Options

        310   Data Validation
        311   Health and Safety
12
12
12
12

12

12
13

13
13

13
14
14
14
16

16

16
16

16

16
16
16
16
16
16
17
17
17

17
17
                                                  IV

-------
 4.0     SOIL SAMPLING AND SURFACE GEOPHYSICS: SOP #2159

        41     Scope and Application                                                              19
        4.2     Method Summary                                                                  19

                4.21    Magnetics                                                                  19
                4.2.2    Electromagnetics                                                           20
                4.2.3    Electrical Resistivity                                                        20
                4.2.4    Seismic                                                                    21
                4.2.5    Ground Penetrating Radar                                                   22

        4.3     Sample Preservation, Containers, Handling and Storage                                23
        4.4     Interferences and Potential Problems                                                 23
        4.5     Equipment/Apparatus                                                              24

                4.5.1    Magnetics                                                                  24
                4.5.2    Electromagnetics                                                           24
                4.5.3    Electrical Resistivity                                                        24
                4.5.4    Seismic                                                                    24
                4.5.5    Ground Penetrating Radar                                                   24

        4.6     Reagents                                                                          24
        4.7     Procedures                                                                         24
        4.8     Calculations                                                                       24
        4.9     Quality Assurance/Quality Control                                                   24
        4.10     Data Validation                                                                   24
        4.11     Health and Safety                                                                 24
APPENDIX A - Figures                                                                           25

APPENDIX B - HNU Field Protocol                                                               29

REFERENCES                                                                                   33

-------
                                       List of Exhibits
Exhibit






Table 1:



Figure 1:




Figure 2:



Figure 3:
                                                  SOP
Recommended Solvent Rinse for Soluble Contaminants   #2006




Sampling Augers                                    #2012




Sampling Trier                                     #2012




Sampling Train Schematic                            #2149
Page






   4




  26



  27




  28
                                               VI

-------
                                    Acknowledgments


Preparation of this document was directed by William A. Coakley, the Removal Program QA Coordinator of
the Environmental Response Team, Emergency Response Division. Additional support was provided under U.S.
EPA contract #68-03-3482 and U.S. EPA contract #68-WO-0036.
                                             VII

-------

-------
     1.0    SAMPLING EQUIPMENT DECONTAMINATION:  SOP #2006
1.1     SCOPE AND APPLICATION

This Standard Operating Procedure (SOP) describes
methods used  for preventing or reducing  cross-
contamination,  and provides general guidelines for
sampling equipment decontamination procedures at
a hazardous waste site.  Preventing or minimizing
cross-contamination  in  sampled media and  in
samples is important for preventing the introduction
of error into sampling results and for protecting the
health and safety of site personnel.

Removing  or neutralizing contaminants that have
accumulated  on  sampling equipment  ensures
protection of personnel from permeating substances,
reduces or eliminates transfer of contaminants to
clean areas,  prevents the mixing of incompatible
substances, and minimizes the likelihood of sample
cross-contamination.
1.2     METHOD SUMMARY

Contaminants  can be  physically removed  from
equipment,  or deactivated  by sterilization  or
disinfection.   Gross contamination of equipment
requires   physical  decontamination,   including
abrasive and non-abrasive methods.  These include
the use of brushes, air and wet blasting, and high-
pressure water cleaning, followed by a wash/rinse
process using appropriate cleaning solutions. Use
of  a  solvent  rinse  is  required when  organic
contamination is present.
 1.3     SAMPLE PRESERVATION,
        CONTAINERS, HANDLING, AND
        STORAGE

 This section is not applicable to this SOP.
 1.4    INTERFERENCES AND
        POTENTIAL PROBLEMS

    •   The  use  of distilled/deionized  water
        commonly  available  from  commercial
        vendors   may   be   acceptable  for
        decontamination of sampling equipment
       provided that it  has been  verified  by
       laboratory analysis to be analyte free.

   •   An untreated potable water supply is not
       an acceptable substitute for tap water. Tap
       water may be used  from  any  municipal
       water  treatment system  for  mixing  of
       decontamination solutions.

   •   Acids   and  solvents  utilized  in  the
       decontamination sequence pose the health
       and safety  risks  of inhalation or skin
       contact,  and raise shipping concerns of
       permeation or degradation.

   •   The site work plan must address disposal
       of the spent decontamination solutions.

   •   Several procedures can be established to
       minimize contact  with  waste and  the
       potential for contamination. For example:

              Stress   work   practices  that
              minimize contact with  hazardous
              substances.

              Use remote sampling, handling,
              and container-opening techniques
              when appropriate.

              Cover monitoring and sampling
              equipment with protective material
              to minimize contamination.

              Use disposable outer garments
              and  disposable   sampling
              equipment when appropriate.
1.5    EQUIPMENT/APPARATUS
       appropriate personal protective clothing
       non-phosphate detergent
       selected solvents
       long-handled brushes
       drop cloths/plastic sheeting
       trash container
       paper towels
       galvanized tubs or buckets
       tap water

-------
         distilled/deionized water
         metal/plastic  containers for  storage  and
         disposal of contaminated wash solutions
         pressurized   sprayers   for   tap   and
         deionized/distilled water
         sprayers for solvents
         trash bags
         aluminum foil
         safety glasses or splash shield
         emergency eyewash bottle
 1.6     REAGENTS

 There are no reagents used in this procedure aside
 from  the  actual  decontamination  solutions and
 solvents.   In  general, the following  solvents  are
 utilized for decontamination purposes:

     •   10% nitric acid(1)
     •   acetone (pesticide grade)(2)
     •   hexane (pesticide grade)(2)
     •   methanol

 (1) Only if sample is to be analyzed for trace metals.
 (2) Only if sample is to be analyzed for organics.
 1.7    PROCEDURES

 As part of the health and safety plan, develop and
 set up a decontamination plan before any personnel
 or equipment enter the areas of potential exposure.
 The  equipment  decontamination   plan  should
 include:

    •   the  number,  location,  and  layout  of
        decontamination stations

    •   which decontamination apparatus is needed

    •   the appropriate decontamination methods

    •   methods  for  disposal of  contaminated
        clothing, apparatus, and solutions

 1.7.1   Decontamination Methods

All personnel, samples, and equipment leaving the
 contaminated   area  of   a  site  must   be
 decontaminated. Various decontamination methods
will   either  physically   remove   contaminants,
inactivate    contaminants   by  disinfection   or
sterilization, or do  both.
 In many cases, gross contamination can be removed
 by physical means.  The physical decontamination
 techniques   appropriate   for  equipment
 decontamination   can  be   grouped  into  two
 categories:   abrasive methods  and non-abrasive
 methods.

 Abrasive Cleaning Methods

 Abrasive cleaning methods work by rubbing and
 wearing away the top layer of the surface containing
 the contaminant.  The following abrasive methods
 are available:

     •  Mechanical: Mechanical cleaning methods
        use  brushes  of metal or nylon.    The
        amount and type of contaminants removed
        will  vary  with  the hardness  of bristles,
        length  of brushing time, and degree of
        brush contact.

     •  Air  Blasting:   Air blasting is  used for
        cleaning   large   equipment,    such  as
        bulldozers,  drilling rigs  or auger bits.  The
        equipment  used  in air  blast  cleaning
        employs compressed  air to force abrasive
        material through a nozzle at high velocities.
        The  distance  between the nozzle and the
        surface cleaned, as  well as the pressure of
        air, the time of  application, and  the angle
        at which  the  abrasive strikes the surface,
        determines cleaning efficiency.  Air blasting
        has several disadvantages:  it is unable to
        control the  amount  of material removed, it
        can aerate contaminants, and it generates
        large amounts of waste.

    •  Wet  Blasting:  Wet  blast cleaning,  also
        used to clean large equipment,  involves use
        of a  suspended  fine abrasive delivered by
        compressed air to the contaminated area.
        The amount of  materials removed can be
        carefully  controlled by using very  fine
        abrasives.  This  method generates a large
        amount of waste.

Non-Abrasive Cleaning Methods

Non-abrasive cleaning  methods work by forcing the
contaminant off of a  surface with pressure. In
general, less of the  equipment surface is removed
using non-abrasive methods.  The following non-
abrasive methods are available:

-------
    •   High-Pressure   Water:    This  method
        consists of  a  high-pressure pump,  an
        operator-controlled directional nozzle, and
        a high pressure hose.  Operating pressure
        usually ranges from 340 to 680 atmospheres
        (atm) which relates to flow rates of 20 to
        140 liters per minute.

    *   Ultra-High-Pressure Water:  This system
        produces a  pressurized water jet  (from
        1,000  to 4,000  atm).   The ultra-high-
        pressure  spray  removes  tightly-adhered
        surface film.   The water velocity ranges
        from 500 m/sec (1,000 atm) to 900  m/sec
        (4,000 atm).  Additives can  enhance the
        method. This method is not applicable for
        hand-held sampling equipment.

Disinfection/Rinse Methods

    •   Disinfection: Disinfectants are a practical
        means of inactivating infectious agents.

    •   Sterilization:      Standard   sterilization
        methods involve heating the equipment.
        Sterilization  is   impractical   for  large
        equipment.

    •   Rinsing:  Rinsing removes contaminants
        through dilution,  physical attraction,  and
        solubilization.

1.7.2   Field Sampling Equipment
        Cleaning  Procedures

Solvent  rinses are  not  necessarily  required when
organics are not a contaminant of concern and may
be eliminated from  the sequence specified  below.
Similarly, an  acid rinse is not required if analysis
does not include inorganics.

1.   Where applicable,  follow  physical  removal
    procedures specified in section L7J..

2.   Wash  equipment   with   a   non-phosphate
    detergent solution.

3.   Rinse with tap water.

4.   Rinse with distilled/deionized water.

5.   Rinse with 10% nitric acid if the sample will be
    analyzed for trace organics.
6.  Rinse with distilled/deionized water.
7.
    Use a solvent rinse (pesticide  grade)  if the
    sample will be analyzed for organics.

8.  Air dry the equipment completely.

9.  Rinse again with distilled/deionized water.

Selection   of   the  solvent   for   use   in  the
decontamination   process   is   based  on  the
contaminants present at the site.  Use of a solvent
is required when organic contamination is present
on-site.    Typical  solvents  used  for  removal  of
organic  contaminants include acetone, hexane, or
water. An acid rinse step is required if metals are
present on-site.  If a particular contaminant fraction
is  not   present   at   the   site,  the nine-step
decontamination procedure listed above  may  be
modified for site specificity.  The decontamination
solvent used should not be among the contaminants
of concern at the site.

Table 1 lists solvent rinses which may be required
for elimination of particular chemicals. After each
solvent rinse, the equipment should be air dried and
rinsed with distilled/deionized water.

Sampling equipment that requires the use of plastic
tubing  should be  disassembled and the tubing
replaced with clean tubing, before commencement
of sampling and between sampling locations.
1.8     CALCULATIONS

This section is not applicable to this SOP.
1.9     QUALITY ASSURANCE/
        QUALITY CONTROL

One type of quality control sample specific to the
field decontamination process is the rinsate blank.
The  rinsate blank  provides  information  on the
effectiveness  of  the  decontamination   process
employed in the field.  When used in conjunction
with field blanks and trip blanks, a rinsate blank can
detect  contamination  during sample  handling,
storage and sample transportation to the laboratory.

-------
             Table 1:  Recommended Solvent Rinse for Soluble Contaminants
                SOLVENT
                                                       SOLUBLE CONTAMINANTS
  Water
•   Low-chain hydrocarbons
•   Inorganic compounds
                                              Salts
                                              Some organic acids and other polar compounds
  Dilute Acids
•   Basic (caustic) compounds
•   Amines
                                             Hydrazines
  Dilute Bases — for example, detergent
  and soap
•   Metals
•   Acidic compounds
•   Phenol
•   Thiols
•   Some nitro and sulfonic compounds
  Organic Solvents1'5 - for example,
  alcohols, ethers, ketones, aromatics,
  straight-chain alkanes (e.g., hexane), and
  common petroleum products (e.g., fuel,
  oil, kerosene)
    Nonpolar compounds (e.g., some organic compounds)
(1) - WARNING: Some organic solvents can permeate and/or degrade protective clothing.
A rinsate blank consists of a sample of analyte-free
(i.c, dcionized) water which  is passed over and
through a field decontaminated sampling device and
placed in a clean sample container.

Rinsate blanks should be run for all parameters of
interest at a rate of 1 per 20  for each parameter,
even if samples are not shipped that day. Rinsate
blanks  are not required  if  dedicated sampling
equipment is used.
1.10    DATA VALIDATION

This section is not applicable to this SOP.


1.11    HEALTH AND SAFETY

When working with potentially hazardous materials,
follow U.S. EPA, OSHA and  specific health  and
safely procedures.

Decontamination can pose hazards under certain
circumstances even though performed to protect
           health and safety.  Hazardous substances may be
           incompatible with decontamination methods. For
           example, the decontamination solution or solvent
           may  react with contaminants  to  produce  heat,
           explosion,  or toxic  products.  Decontamination
           methods may be incompatible with  clothing or
           equipment; some solvents can  permeate or degrade
           protective clothing. Also, decontamination solutions
           and solvents  may pose a direct health hazard to
           workers through inhalation or skin contact, or if
           they combust.

           The decontamination solutions and solvents must be
           determined  to  be compatible before  use.   Any
           method  that permeates, degrades, or  damages
           personal protective equipment should not be used.
           If decontamination methods pose  a direct  health
           hazard, measures  should  be  taken  to  protect
           personnel or the methods should  be modified to
           eliminate the hazard.

-------
                         2.0    SOIL SAMPLING:  SOP #2012
 2.1    SCOPE AND APPLICATION

 The purpose of this Standard Operating Procedure
 (SOP) is to describe the procedures for collecting
 representative soil samples. Analysis of soil samples
 may determine whether concentrations of specific
 soil pollutants exceed established action levels, or if
 the concentrations of soil pollutants present a risk
 to public health, welfare, or the environment.
 2.2    METHOD SUMMARY

 Soil samples may be collected using a variety of
 methods and  equipment.    The  methods  and
 equipment used are dependent on the depth of the
 desired sample, the  type  of  sample required
 (disturbed versus undisturbed), and the type of soil.
 Near-surface soils may be easily sampled using a
 spade, trowel,  and scoop.   Sampling at greater
 depths may be performed using a hand auger, a
 trier, a split-spoon, or, if required, a backhoe.
2.3    SAMPLE PRESERVATION,
        CONTAINERS, HANDLING, AND
        STORAGE

Chemical preservation of solids  is not generally
recommended. Refrigeration to 4°C, supplemented
by  a  minimal holding time,  is usually the best
approach.
2.4    INTERFERENCES AND
       POTENTIAL PROBLEMS

There are two primary interferences or potential
problems associated with  soil sampling.  These
include  cross-contamination   of  samples  and
improper sample collection.  Cross-contamination
problems can be eliminated or minimized through
the use of dedicated sampling equipment.  If this is
not possible or practical, then decontamination of
sampling equipment is necessary. Improper sample
collection   can  involve   using   contaminated
equipment, disturbance of  the matrix resulting in
compaction  of  the  sample,  or   inadequate
homogenization of  the samples where  required,
 resulting in variable, non-representative results.
 2.5    EQUIPMENT/APPARATUS
        sampling plan
        maps/plot plan
        safety equipment, as specified in the health
        and safety plan
        compass
        tape measure
        survey stakes or flags
        camera and film
        stainless steel, plastic, or other appropriate
        homogenization bucket or bowl
        1-quart mason jars w/Teflon liners
        Ziploc plastic bags
        logbook
        labels
        chain of custody forms and seals
        field data sheets
        cooler(s)
        ice
        decontamination supplies/equipment
        canvas or plastic sheet
        spade or shovel
        spatula
        scoop
        plastic or stainless steel spoons
        trowel
        continuous flight (screw) auger
        bucket auger
        post hole auger
        extension rods
        T-handle
        sampling trier
        thin-wall tube sampler
        Vehimeyer soil sampler outfit
        - tubes
        - points
        - drive head
        - drop hammer
        - puller jack and grip
        backhoe
2.6    REAGENTS

Reagents are not used for the preservation of soil
samples. Decontamination solutions are specified in

-------
ERT   SOP   #2006,   Sampling   Equipment
Decontamination.
2.7    PROCEDURES

2.7.1  Preparation

1.  Determine the extent of the sampling effort, the
    sampling methods to be employed, and which
    equipment and supplies are required.

2.  Obtain  necessary sampling  and monitoring
    equipment.

3.  Decontaminate or preclean  equipment,  and
    ensure that it is in working order.

4.  Prepare schedules, and coordinate with staff,
    client, and regulatory agencies, if appropriate.

5.  Perform a general site survey prior to site entry
    in accordance with the site-specific health and
    safety plan.

6.  Use stakes, buoys, or flagging to identify and
    mark all sampling locations.  Consider specific
    site factors, including extent  and nature of
    contaminant, when selecting sample location. If
    required,  the proposed  locations  may  be
    adjusted   based   on   site access,  property
    boundaries,  and  surface  obstructions.   All
    staked locations  will be utility-cleared by the
    property owner prior to soil sampling.

2.7.2  Sample Collection

Surface Soil Samples

Collect samples from near-surface soil with tools
such  as spades,  shovels,  and scoops.   Surface
material can be removed to the required depth with
this equipment,  then a stainless steel or plastic
scoop can be used to collect the sample.

This method can be used in most soil types but is
limited to sampling near surface areas. Accurate,
representative  samples can be collected with  this
procedure depending on  the  care  and  precision
demonstrated by the sampling team member. The
use of a flat, pointed mason trowel to cut a block of
the desired soil can be helpful when undisturbed
profiles are required.  A stainless steel scoop, lab
spoon, or plastic spoon will suffice in most other
applications.  Avoid the use of devices plated with
chrome or other materials. Plating is particularly
common with garden implements such as potting
trowels.

Follow  these  procedures to  collect surface  soil
samples.

1.  Carefully remove the top layer of soil or debris
    to the desired sample depth with a pre-cleaned
    spade.

2.  Using a  pre-cleaned,  stainless  steel  scoop,
    plastic spoon, or trowel, remove and discard a
    thin layer of soil from the area which came in
    contact with the spade.

3.  If volatile organic analysis is to be performed,
    transfer a portion of the sample directly into an
    appropriate, labeled sample container(s) with a
    stainless steel lab spoon,  plastic lab spoon, or
    equivalent and secure the cap(s) tightly. Place
    the remainder of the sample into a  stainless
    steel,   plastic,   or  other   appropriate
    homogenization container, and mix thoroughly
    to obtain a homogenous sample representative
    of the entire sampling interval. Then, either
    place the sample into an appropriate, labeled
    container(s) and secure the cap(s) tightly; or, if
    composite samples are to be collected, place a
    sample from another sampling interval into the
    homogenization container and mix thoroughly.
    When compositing  is complete,  place  the
    sample into appropriate,  labeled container(s)
    and secure the cap(s) tightly.

Sampling at Depth  with Augers and Thin-
Wall Tube Samplers

This  system  consists of an  auger,  a series  of
extensions, a  "T"  handle,  and a  thin-wall tube
sampler (Appendix A, Figure 1). The auger is used
to bore a hole to a desired  sampling depth, and is
then withdrawn.  The sample may be  collected
directly from the auger. If a core sample is to be
collected, the auger tip is then replaced with a thin-
wall tube  sampler.   The system is  then  lowered
down the borehole, and driven into the soil at the
completion depth. The system is withdrawn and the
core collected from the thin-wall tube sampler.

Several  types  of augers  are available.    These
include:    bucket, continuous flight (screw), and
pesthole augers. Bucket augers are better for direct

-------
sample recovery since they provide a large volume
of sample in a short time.  When continuous flight
augers  are used, the  sample  can be  collected
directly from the flights, which are usually at 5-feet
intervals.    The  continuous  flight  augers  are
satisfactory for use when  a  composite  of  the
complete soil column is  desired.  Pesthole augers
have limited utility for sample collection as they are
designed to cut through fibrous,  rooted, swampy
soil.

Follow these procedures for collecting soil samples
with the auger and a thin-wall tube sampler.

1.  Attach the auger bit to  a drill rod extension,
    and attach the "T" handle to the drill rod.

2.  Clear the area to be sampled of any surface
    debris  (e.g., twigs, rocks, litter).  It may be
    advisable  to remove  the first 3 to 6  inches of
    surface soil for an area approximately 6 inches
    in radius around the drilling location.

3.  Begin  augering, periodically removing  and
    depositing accumulated  soils onto  a plastic
    sheet  spread  near  the hole.   This prevents
    accidental brushing of loose material back down
    the borehole  when  removing the  auger or
    adding drill rods. It also facilitates refilling the
    hole, and avoids possible contamination of the
    surrounding area.

4.  After  reaching the desired depth, slowly  and
    carefully remove the auger from boring. When
    sampling directly from the auger, collect sample
    after the  auger  is removed from boring  and
    proceed to Step  10.

5.  Remove auger tip from drill rods and replace
    with a  pre-cleaned  thin-wall tube  sampler.
    Install proper cutting tip.

6.  Carefully  lower  the  tube sampler  down  the
    borehole.   Gradually force  the tube sampler
    into the soil.  Care should be taken to avoid
    scraping the borehole sides. Avoid hammering
    the drill  rods  to  facilitate  coring as  the
    vibrations may  cause   the  boring  walls  to
    collapse.

7.  Remove the tube sampler, and unscrew the drill
    rods.

8.  Remove the cutting tip and the core from the
    device.
9.  Discard the top of the core (approximately 1
    inch), as this represents material collected
    before  penetration of  the  layer of concern.
    Place the remaining core into the appropriate
    labeled   sample   container(s).     Sample
    homogenization is not required.

10. If volatile organic analysis is to be performed,
    transfer a portion of the sample directly into an
    appropriate, labeled sample containers) with a
    stainless steel lab spoon, plastic lab spoon, or
    equivalent and secure the cap(s) tightly. Place
    the remainder of the sample into  a stainless
    steel,  plastic,    or  other   appropriate
    homogenization container, and mix thoroughly
    to obtain a homogenous sample representative
    of the entire sampling interval.  Then, either
    place the sample into an appropriate, labeled
    container(s) and secure the cap(s) tightly; or, if
    composite samples are to be collected, place a
    sample from another sampling interval into the
    homogenization container and mix thoroughly.
    When  compositing is  complete,  place  the
    sample   into   the   appropriate,   labeled
    container(s) and secure the cap(s) tightly.

11. If another sample is to be collected in the same
    hole,  but at a greater depth, reattach the auger
    bit to the drill and assembly, and follow steps
    3 through 11, making sure  to decontaminate
    the auger and tube sampler between samples.

12. Abandon the hole according to applicable state
    regulations.    Generally,  shallow  holes  can
    simply be backfilled with the  removed  soil
    material.

Sampling at Depth with a  Trier

The system consists of a trier, and a "T" handle.
The auger is driven into the soil to be sampled and
used to extract a core sample from the appropriate
depth.

Follow these procedures to collect soil samples with
a sampling trier.

1.  Insert the trier (Appendix A, Figure 2) into the
    material to be sampled at a 0° to 45° angle
    from  horizontal.   This  orientation  minimizes
    the spillage of sample.

2.  Rotate the trier once  or twice to cut a core of
    material.

-------
3.  Slowly withdraw the trier, making sure that the
    slot is faring upward.

4.  If volatile organic analysis is to be performed,
    transfer a portion of the sample directly into an
    appropriate, labeled sample containers) with a
    stainless steel lab spoon, plastic lab spoon, or
    equivalent and secure the cap(s) tightly.  Place
    the remainder of the sample  into a stainless
    steel,   plastic,   or   other   appropriate
    homogenization container, and mix thoroughly
    to obtain a homogenous sample representative
    of the entire sampling interval.  Then, either
    place the sample into an appropriate, labeled
    container^) and secure the cap(s)  tightly; or, if
    composite samples are  to be collected, place a
    sample from another sampling interval into the
    homogenization container and  mix thoroughly.
    When  compositing  is  complete, place  the
    sample into an appropriate, labeled container(s)
    and secure the cap(s) tightly.

Sampling  at Depth  with  a  Split Spoon
(Barrel) Sampler

The procedure for split spoon sampling describes
the collection and extraction  of undisturbed  soil
cores of 18 or 24 inches in length.   A series of
consecutive  cores may be extracted  with a split
spoon sampler to give  a  complete  soil  column
profile, or an auger may be used to drill down to
the desired depth for sampling. The split spoon is
then  driven  to its  sampling  depth through  the
bottom of the augured hole and the core extracted.

When split  tube  sampling is performed to gain
geologic information, all work should be performed
in accordance with ASTM D 1586-67  (reapproved
1974).

Follow these procedures for collecting soil samples
with a split spoon.

1.  Assemble the sampler by aligning both sides of
    the barrel and then screwing the  bit onto the
    bottom  and the heavier head piece onto the
    top.

2.  Place the sampler in a perpendicular position
    on the sample material.

3.  Using  a  sledge  hammer or well ring, if
    available, drive the tube. Do not drive past the
    bottom of the head piece or compression  of the
    sample will result.

4.   Record in  the site  logbook or on field data
    sheets the length of the tube used to penetrate
    the material being sampled, and the number of
    blows required to obtain this depth.

5.   Withdraw the sampler, and open by unscrewing
    the bit and head and splitting the barrel.  If a
    split sample is desired, a cleaned, stainless steel
    knife should be used to divide the tube contents
    in hall', longitudinally.  This sampler is typically
    available in diameters of 2 and 3 1/2 niches.
    However, in order to obtain the  required
    sample volume, use of a larger barrel may be
    required.

6.   Without disturbing  the core, transfer it to an
    appropriate labeled sample containers)  and
    seal tightly.

Test Pit/Trench Excavation

These relatively  large  excavations are  used to
remove  sections of soil, when detailed examination
of soil characteristics (horizontal structure, color,
etc.) are required.   It  is  the  least cost effective
sampling method due to the relatively high cost of
backhoe operation.

Follow these procedures for collecting soil samples
from test pit/trench excavations.

1.   Prior  to  any excavation with a backhoe,  it is
    important to ensure that all sampling locations
    are clear of utility lines and poles (subsurface
    as well as above surface).

2.   Using   the  backhoe,  dig   a   trench  to
    approximately   3   feet   in   width   and
    approximately  1  foot below  the cleared
    sampling location. Place removed or excavated
    soils on plastic sheets.  Trenches greater than
    5 feet deep must be sloped or protected by a
    shoring  system,   as   required   by  OSHA
    regulations.

3.   Use a shovel to remove a 1- to 2-inch layer of
    soil from the vertical  face of the pit  where
    sampling is to be done.

4.   Take samples using a trowel, scoop, or  coring
    device at the desired  intervals.   Be sure to
    scrape the vertical face at the point of sampling

-------
    to remove any soil that may have fallen from
    above, and to expose fresh soil for sampling. In
    many instances, samples  can be collected
    directly from the backhoe bucket.

5.   If volatile organic analysis is to be performed,
    transfer a portion of the sample directly into an
    appropriate, labeled sample container(s) with a
    stainless steel lab spoon, plastic lab spoon, or
    equivalent and secure the cap(s) tightly.  Place
    the remainder of  the sample  into a stainless
    steel,   plastic,  or   other   appropriate
    homogenization container, and mix thoroughly
    to obtain a homogenous sample representative
    of the entire  sampling interval.  Then, either
    place the sample into an appropriate, labeled
    container(s) and secure the  cap(s) tightly; or, if
    composite samples are to be collected, place a
    sample from another sampling interval into the
    homogenization container and mix thoroughly.
    When  compositing  is  complete,  place the
    sample into appropriate, labeled container(s)
    and secure the cap(s) tightly.

6.  Abandon the  pit  or  excavation according to
    applicable state regulations. Generally, shallow
    excavations can simply be  backfilled with the
    removed soil material.
2.8    CALCULATIONS

This section is not applicable to this SOP.
2.9    QUALITY ASSURANCE/
       QUALITY CONTROL

There are no  specific quality assurance  activities
which  apply  to  the implementation  of  these
procedures. However, the following QA procedures
apply:

    •  All data must be documented on field data
       sheets or within site logbooks.

    •  All instrumentation must be operated in
       accordance with operating instructions as
       supplied  by  the manufacturer,  unless
       otherwise  specified  in  the work  plan.
       Equipment   checkout   and  calibration
       activities   must   occur   prior   to
       sampling/operation,  and they  must  be
       documented.
2.10   DATA VALIDATION

This section is not applicable to this SOP.


2.11   HEALTH AND SAFETY

When working with potentially hazardous materials,
follow U.S.  EPA, OSHA, and specific health and
safety procedures.

-------

-------
                     3.0    SOIL GAS SAMPLING:  SOP #2149
3.1     SCOPE AND APPLICATION

Soil gas monitoring provides a quick means of waste
site evaluation.  Using this method, underground
contamination can be identified,  and the source,
extent, and  movement of the pollutants  can  be
traced.

This Standard Operating Procedure (SOP) outlines
the methods used by EPA/ERT in installing soil gas
wells'; measuring organic levels in the soil gas using
an HNU PI 101 Portable Photoionization Analyzer
and/or other air monitoring devices; and sampling
the soil gas using Tedlar bags, Tenax sorbent tubes,
and SUMMA canisters.
3.2    METHOD SUMMARY

A 3/8-inch diameter hole is driven into the ground
to a depth of 4 to 5 feet using a commercially
available "slam bar". (Soil gas can also be sampled
at other depths by the use of a longer bar or bar
attachments.) A 1/4-inch O.D. stainless steel probe
is inserted into the hole. The hole is then sealed at
the  top around the probe using modeling clay. The
gas  contained in the interstitial spaces of the soil is
sampled by pulling the sample through the probe
using an air  sampling pump.  The sample may be
stored  in  Tedlar  bags,  drawn  through sorbent
cartridges, or  analyzed  directly using  a  direct
reading instrument.

The air sampling pump is not used for SUMMA
canister sampling of soil gas. Sampling is achieved
by  soil gas equilibration  with the  evacuated
SUMMA  canister.  Other  field air monitoring
devices, such as the combustible gas indicator (MSA
CGI/02 Meter,  Model 260) and the organic vapor
analyzer (Foxboro OVA, Model  128), can also be
used  depending  on  specific  site  conditions.
Measurement   of  soil   temperature   using  a
temperature  probe  may also be desirable. Bagged
samples are  usually analyzed in a field  laboratory
using a portable Photovac GC.

Power driven sampling probes may be utilized when
soil conditions make sampling by hand unfeasible
(i.e., frozen ground, very dense clays, pavement,
etc.).   Commercially available soil gas sampling
probes (hollow, 1/2-inch O.D. steel probes) can be
driven to the desired depth using a power hammer
(e.g., Bosch Demolition Hammer).  Samples can be
drawn through the probe itself, or through Teflon
tubing inserted through the probe and attached to
the probe point.   Samples  are collected  and
analyzed as described above.
3.3    SAMPLE PRESERVATION,
        CONTAINERS, HANDLING, AND
        STORAGE

3.3.1   Tedlar Bag

Soil gas samples are generally contained in  1-L
Tedlar bags.  Bagged samples are best stored in
coolers to protect the bags from any damage that
may occur in the field or in  transit.  In addition,
coolers  ensure  the integrity of the samples  by
keeping them at a cool temperature and out of
direct sunlight. Samples should be analyzed as soon
as possible, preferably within 24 to 48 hours.

3.3.2  Tenax Tube

Bagged  samples can also be drawn into Tenax or
other sorbent tubes to undergo lab GC/MS analysis.
If Tenax tubes are to be utilized, special care must
be taken to avoid contamination. Handling of the
tubes  should be kept to a minimum, and samplers
must wear nylon or other lint-free gloves.  After
sampling, each tube should be stored in a clean,
sealed culture tube; the ends packed  with clean
glass  wool  to  protect  the  sorbent  tube from
breakage.  The  culture tubes should be kept cool
and wrapped in aluminum  foil to  prevent  any
photodegradation of samples (see Section 3.7.4.).

3.3.3  SUMMA Canister

The SUMMA canisters used for soil gas sampling
have a 6-L sample capacity and are certified clean
by GC/MS  analysis before being utilized in  the
field.  After sampling is completed, they are stored
and shipped  in travel cases.
                                               11

-------
 3.4    INTERFERENCES AND
        POTENTIAL PROBLEMS

 3.4.1  HNU Measurements

 A number of factors can affect the response of the
 HNU PI 101.   High  humidity can  cause  lamp
 fogging and decreased sensitivity.  This  can be
 significant when soil moisture levels are high, or
 when a soil gas well is actually in groundwater.
 High concentrations  of  methane can cause  a
 downscale deflection of the meter. High and low
 temperature,   electrical   fields,  FM   radio
 transmission, and naturally occurring  compounds,
 such as terpenes in wooded areas, will also affect
 instrument response.

 Other field screening instruments can be affected by
 interferences. Consult the manufacturers' manuals.

 3.4.2  Factors Affecting Organic
        Concentrations in Soil Gas

 Concentrations   in  soil  gas  are  affected  by
 dissolution,   adsorption,  and   partitioning.
 Partitioning refers to the ratio of component found
 in a saturated vapor above an aqueous solution to
 the amount in the solution; this can, in theory, be
 calculated  using  the  Henry's  Law  constants.
 Contaminants can also be adsorbed onto inorganic
 soil   components  or  "dissolved"  in  organic
 components. These factors can result in a lowering
 of the partitioning coefficient.

 Soil "tightness" or amount of void space hi the soil
 matrix, will affect the rate of recharging of gas into
 the soil gas well.

 Existence of a high, or perched, water table, or of
 an impermeable underlying layer (such as a  clay
 lens or layer of buried slag) may interfere with
 sampling of the soil gas. Knowledge of site geology
 is useful in  such situations,  and can prevent
 inaccurate sampling.

 3.4.3  Soil Probe Clogging

A common problem with  this sampling method is
 soil probe  clogging.   A  clogged  probe can be
 identified by using an in-line vacuum gauge or by
 listening for the sound of the pump laboring.  This
problem can usually be eliminated by using a wire
 cable  to clear the probe  (see procedure #3 in
Section 3.7.1).
3.4.4  Underground Utilities

Prior to selecting sample locations, an underground
utility search is recommended.  The local utility
companies can be contacted and requested to mark
the locations of their underground lines. Sampling
plans  can then be drawn up  accordingly.   Each
sample  location should also be  screened with a
metal detector  or  magnetometer to verify that no
underground pipes or drums exist.
3.5    EQUIPMENT/APPARATUS

3.5.1  Slam Bar Method
       slam bar (one per sampling team)
       soil gas probes, stainless steel tubing, 1/4-
       inch O.D., 5 foot length
       flexible wire or cable used for clearing the
       tubing during insertion into the well
       "quick connect" fittings to connect sampling
       probe tubing, monitoring instruments, and
       Gilian pumps to  appropriate fittings  on
       vacuum box
       modeling clay
       va.cuum box for drawing a vacuum around
       Tedlar bag for sample  collection (one per
       sampling team)
       Gilian pump Model HFS113A adjusted to
       approximately 3.0  L/min (one to two per
       sampling team)
       1/4-inch Teflon tubing, 2 to 3 foot lengths,
       for replacement of contaminated sample
       line
       Tedlar bags, 1 liter, at least one bag per
       sample point
       soil gas sampling labels, field data sheets,
       logbook, etc.
       HNU  Model PI  101,  or  other field air
       monitoring devices, (one  per sampling
       team)
       ice chest, for carrying  equipment and for
       protection of samples  (two per sampling
       team)
       metal  detector   or magnetometer, for
       detecting  underground   utilities/
       pipes/drums (one  per sampling team)
       Photovac  GC, for field-lab  analysis  of
       bagged samples
       SUMMA  canisters (plus  then*  shipping
       cases)   for   sample,   storage   and
       transportation
                                               12

-------
3.5.2  Power Hammer Method

    •   Bosch demolition hammer
    •   1/2-inch O.D. steel probes, extensions, and
        points
    •   dedicated aluminum sampling points
    •   Teflon tubing, 1/4-inch O.D.
    •   "quick connect" fittings to connect sampling
        probe tubing, monitoring instruments, and
        Gilian pumps to  appropriate fittings on
        vacuum box
    •   modeling clay
    •   vacuum box for drawing a vacuum around
        Tedlar bag for sample collection (one per
        sampling team)
    •   Gilian pump Model HFS113A adjusted to
        approximately 3.0  L/min (one to two per
        sampling team)
    •   1/4-inch Teflon tubing, 2 to 3 foot lengths,
        for replacement of contaminated sample
        line
    •   Tedlar bags, 1 liter, at least one bag per
        sample point
    •   soil gas sampling labels, field data sheets,
        logbook, etc.
    •   HNU Model PI 101,  or  other field air
        monitoring  devices, (one  per  sampling
        team)
    •   ice chest, for carrying equipment and for
        protection of samples (two per  sampling
        team)
    •   metal detector   or magnetometer,  for
        detecting   underground   utilities/
        pipes/drums (one per sampling team)
    •   Photovac GC,  for field-lab  analysis  of
        bagged samples
    •   SUMMA canisters (plus their  shipping
        cases)  for   sample,   storage   and
        transportation
    •   generator with extension cords
    •   high lift jack assembly for removing probes
3.6     REAGENTS

    •   HNU  Systems Inc.  Calibration Gas  for
        HNU Model PI 101, and/or calibration gas
        for other field air monitoring devices
    •   deionized  organic-free   water,   for
        decontamination
    •   methanol,   HPLC   grade,   for
        decontamination
    •   ultra-zero grade compressed air, for field
        blanks
        standard gas preparations for Photovac GC
        calibration and Tedlar bag spikes
3.7    PROCEDURES

3.7.1   Soil Gas Well Installation

1.  Initially, make a hole slightly deeper than the
    desired depth. For sampling up to 5 feet, use
    a 5-foot single piston slam bar.  For deeper
    depths, use a piston slam bar with threaded 4-
    foot-long extensions.  Other techniques can be
    used, so long as holes are of narrow diameter
    and no contamination is introduced.

2.  After the hole is made, carefully withdraw the
    slam bar to prevent collapse of the walls of the
    hole. Then insert the soil gas probe.

3.  It is necessary to prevent plugging of the probe,
    especially for deeper holes. Place a metal wire
    or cable, slightly longer than the probe, into the
    probe prior to inserting into the hole.  Insert
    the probe to full depth, then pull it up 3 to 6
    inches, then clear it by moving the cable up and
    down. The cable is removed before sampling.

4.  Seal the top of the sample hole at the surface
    against  ambient  air infiltration  by  using
    modeling clay molded around the probe at the
    surface of the hole.

5.  If conditions preclude hand installation of the
    soil gas wells, the power driven  system may be
    employed.     Use   the   generator-powered
    demolition hammer to drive the probe to the
    desired depth (up to 12 feet may be attained
    with extensions).  Pull the probe up 1  to 3
    inches if the retractable point is used. No clay
    is needed to  seal the hole.  After sampling,
    retrieve  the  probe  using  the  high lift  jack
    assembly.

6.  If semi-permanent soil gas wells are required,
    use the  dedicated  aluminum  probe points.
    Insert these  points  into  the bottom of the
    power-driven probe and attach it to the Teflon
    tubing. Insert the probe as in  step 5.  When
    the probe is  removed, the point and Teflon
    tube remain in the hole, which may be sealed
    by backfilling with sand, bentonite, or soil.
                                                13

-------
 3.7.2  Screening with Field
        Instruments

 1.  The well volume must be evacuated prior to
    sampling. Connect the Gilian pump, adjusted
    to 3.0 L/min,  to the sample probe using a
    section of Teflon tubing as a connector.  Turn
    the pump on, and a vacuum is pulled through
    the probe for approximately 15 seconds.  A
    longer tune  is  required for  sample wells of
    greater depths.

 2.  After  evacuation,  connect  the  monitoring
    instruments) to the probe  using  a Teflon
    connector.   When  the  reading is stable, or
    peaks, record  the  reading.   For  detailed
    procedures   on  HNU  field protocol,  see
    appendix B,  and refer to  the manufacturer's
    instructions.

 3.  Some  readings  may be above or below the
    range set on  the field instruments. The range
    may be reset, or the response recorded as a
    figure  greater than  or  less  than the range.
    Consider the  recharge rate of the well with soil
    gas when sampling at a different range setting.

 3.7.3  Tedlar  Bag Sampling

 1.  Follow step 1 in section 3.7.2 to evacuate well
    volume. If air monitoring instrument screening
    was performed prior to sampling, evacuation is
    not necessary.

 2.  Use the vacuum box and sampling train (Figure
    3 in  Appendix A) to take  the sample.  The
    sampling train  is designed to minimize the
    introduction of contaminants and losses due to
    adsorption. All wetted parts are either Teflon
    or stainless  steel.   The  vacuum is  drawn
    indirectly to avoid contamination from sample
    pumps.

3.  Place the Tedlar bag inside the vacuum box,
    and attach it to the sampling port.  Attach the
    sample probe to the sampling port via Teflon
    tubing and a "quick connect" fitting.

4.  Draw a vacuum  around the outside of the bag,
    using a Gilian pump connected to  the vacuum
    box evacuation port, via Tygon tubing and a
    "quick connect" fitting. The vacuum causes the
    bag to inflate, drawing the sample.
5.  Break, the vacuum by removing the Tygon line
    from the pump.  Remove the bagged sample
    from the  box and  close valve.   Label bag,
    record data on data  sheets or in logbooks.
    Record the date, time, sample location ID, and
    the HNU,  or other instrument reading(s) on
    sample bag label.

CAUTION: Labels  should not be pasted directly
onto the bags, nor should bags be labeled directly
using a marker or pen.  Inks and adhesive may
diffuse through the bag material, contaminating the
sample. Place labels  on the edge of the bags, or tie
the labels to the metal eyelets provided on the bags.
Markers with inks containing volatile organics (i.e.,
permanent: ink markers) should not be  used.

3.7.4  Tenax Tube Sampling

Samples collected in Tedlar bags may be sorbed
onto Tenax tubes for further analysis by GC/MS.

Additional Apparatus

    •   Syringe  with a  luer-lock tip  capable of
        drawing a soil gas or air sample from a
        Tedlar  bag  onto  a  Tenax/CMS  sorbent
        tube.  The syringe capacity is dependent
        upon the volume of sample being drawn
        onto the sorbent tube.

    •   Adapters  for  fitting the  sorbent  tube
        between the  Tedlar bag and the sampling
        syringe.  The adapter attaching the Tedlar
        bag to  the  sorbent  tube  consists of a
        reducing union (1/4-inch to 1/16-inch O.D.
        ~  Swagelok cat.  #  SS-400-6-ILV or
        equivalent) with a length of 1/4-inch O.D.
        Teflon tubing replacing the nut  on  the 1/6-
        inch (Tedlar bag)  side.   A 1/4-inch I.D.
        silicone O-ring replaces the ferrules in the
        nut on the 1/4-inch (sorbent tube) side of
        the union.

        The adapter attaching the sampling syringe
        to the sorbent tube consists of a reducing
        union  (1/4-inch to  1/16-inch  O.D. —
        Swagelok  Cat.   #   SS-400-6-ILV  or
        equivalent) with a  1/4-inch I.D.  silicone
        O-ring replacing the ferrules in the nut on
        the 1/4-inch (sorbent tube) side  and the
        needle  of  a luer-lock  syringe   needle
        inserted into the 1/16-inch side  (held in
        place with  a 1/16-inch  ferrule).   The
                                                14

-------
        luer-lock end of the needle can be attached
        to the sampling syringe. It is useful to have
        a luer-lock on/off valve situated between
        the syringe and the needle.

     •  Two-stage glass sampling cartridge (1/4-
        inch O.D. x  1/8-inch I.D. x 5 1/8 inch)
        contained   in   a  flame-sealed  tube
        (manufactured   by   Supelco    Custom
        Tenax/Spherocarb Tubes or equivalent)
        containing two sorbent sections retained by
        glass wool:

        Front section:   150 mg of Tenax-GC
        Back  section:    150 mg of CMS
        (Carbonized Molecular Sieve)

        Sorbent tubes may also be prepared in the
        lab and  stored in either Teflon-capped
        culture  tubes  or  stainless  steel tube
        containers.  Sorbent tubes stored  in this
        manner should not be kept more  than 2
        weeks without reconditioning.   (See SOP
        #2052  for   Tenax/CMS   sorbent tube
        preparation).

     •   Teflon-capped culture tubes  or stainless
        steel  tube containers  for  sorbent tube
        storage.   These  containers should  be
        conditioned by baking at 120°C for at least
        2 hours. The culture tubes should contain
        a glass wool plug to prevent sorbent tube
        breakage during transport.  Reconditioning
        of  the containers  should  occur between
        usage or after extended  periods of disuse
        (i.e., 2 weeks  or more).

     •   Nylon gloves  or lint-free cloth.  (Hewlett
        Packard Part  # 8650-0030 or equivalent.)

Sample Collection

1.  Handle  sorbent tubes with care, using nylon
    gloves (or other  lint-free material) to avoid
    contamination.

2.  Immediately before sampling, break one end of
    the sealed  tube  and  remove  the   Tenax
    cartridge. For in-house prepared tubes, remove
    cartridge from  its  container.

3.  Connect the valve on the Tedlar bag to the
    sorbent tube adapter. Connect the sorbent tube
    to  the sorbent tube  adapter with the  Tenax
 4.
(white granular) side of the tube facing the
Tedlar bag.

Connect the sampling syringe assembly to the
CMS (black) side of the sorbent tube. Fittings
on the adapters should be very tight.
 5.   Open the valve on the Tedlar bag.

 6.   Open the on/off valve of the sampling syringe.

 7.   Draw a predetermined volume of sample onto
     the sorbent tube. (This may require closing the
     syringe valve, emptying the syringe and then
     repeating the procedure, depending upon  the
     syringe  capacity  and   volume  of  sample
     required.)

 8.   After  sampling, remove the tube  from  the
     sampling train with gloves or a clean cloth.  Do
     not label or write on the Tenax/CMS tube.

 9.   Place  the  sorbent  tube  hi  a  conditioned
     stainless steel tube  holder or culture  tube.
     Culture tube caps should be sealed with Teflon
     tape.

 Sample Labeling

 Each sample tube container (not tube)  must be
 labeled with the site name, sample station number,
 sample date, and sample volume.

 Chain of custody forms must accompany all samples
 to the laboratory.

 Quality Assurance

 Before field use, a QA  check should be performed
 on each batch of sorbent tubes by analyzing a tube
 with   thermal   desorption/cryogenic   trapping
 GC/MS.

At least one blank sample must be submitted with
 each set  of samples  collected at a site. This trip
blank must be treated the same as the sample tubes
 except no sample will be drawn through the tube.

Sample tubes should be stored out of UV light (i.e.,
sunlight)  and kept on ice until analysis.

Samples  should  be taken  in  duplicate, when
possible.
                                                 15

-------
3.7.5  SUMMA Canister Sampling

1.   Follow item 1 in step 3.7.2 to evacuate well
    volume. If HNU analysis was performed prior
    to taking a sample, evacuation is not necessary.

2.   Attach  a certified  clean,  evacuated   6-L
    SUMMA  canister via  the 1/4-inch  Teflon
    tubing.

3.   Open the  valve on SUMMA canister. The soil
    gas  sample is drawn  into the  canister by
    pressure  equilibration.    The  approximate
    sampling tune for a 6-L canister is 20 minutes.

4.   Site name, sample location, number, and date
    must be recorded on a chain of custody form
    and on a blank tag attached to the canister.
3.8    CALCULATIONS

3.8.1   Field Screening Instruments

Instrument readings are usually read directly from
the meter. In some cases, the background level at
the soil gas station may be subtracted:
    Final Reading
Sample Reading -
Background
3.8.2  Photovac GC Analysis

Calculations used to determine concentrations of
individual components by Photovac GC analysis are
beyond the scope of this SOP and are covered hi
ERT SOP #2109, Photovac GC Analysis for Soil,
Water and Air/Soil Gas.
 3.9    QUALITY ASSURANCE/
        QUALITY CONTROL

 3.9.1  Field Instrument Calibration

 Consult the manufacturers' manuals for correct use
 and calibration of all instrumentation.  The HNU
 should be calibrated at least once a day.

 3.9.2  Gilian Model  HFS113A Air
        Sampling Pump Calibration

 Flow should  be set at approximately 3.0 L/min;
                            accurate flow adjustment is not necessary.  Pumps
                            should be calibrated prior to bringing into the field.
3.9.3  Sample Probe Contamination

Sample probe contamination is  checked between
each sample by drawing ambient air through the
probe via a Gilian pump and checking the response
of the HNU PI 101.  If HNU readings are higher
than background, replacement or decontamination
is necessary.

Sample probes may be decontaminated simply by
drawing ambient air through the probe until the
HNU reading is at background.  More persistent
contamination can be washed out using methanol
and water, then air drying. Having more than one
probe per  sample team will reduce lag  times
between  sample  stations  while  probes  are
decontaminated.

3.9.4  Sample Train Contamination

The Teflon line forming the sample train from the
probe to the Tedlar  bag  should be changed on a
daily basis.  If visible contamination (soil or water)
is  drawn into the sampling  train, it should be
changed immediately.  When sampling in  highly
contaminated areas,  the sampling train should be
purged with ambient air, via a Gilian pump, for
approximately 30  seconds between each sample.
After purging, the sampling train can be checked
using an HNU, or other field monitoring device, to
establish the cleanliness of the Teflon line.

3.9.5  Field Blank

Each cooler containing samples should also contain
one Tedlar bag of ultra-zero grade air, acting as a
field blank.  The field blank should accompany the
samples in the field (while being collected) and
when they are delivered for analysis. A fresh blank
must be provided to  be placed hi the empty cooler
pending additional sample collection: One new field
blank per cooler of samples is required. A chain of
custody form  must  accompany each cooler of
samples and should include  the  blank  that  is
dedicated to that group of samples.

3.9.6  Trip Standard

Each cooler containing samples should contain a
Tedlar  bag  of  standard  gas  to!  calibrate the
                                               16

-------
 analytical instruments (Photovac GC, etc.).  This
 trip standard will be used to determine any changes
 in concentrations of the target compounds during
 the course  of the  sampling  day (e.g., migration
 through   the  sample   bag,   degradation,   or
 adsorption). A fresh trip standard must be provided
 and placed in each cooler pending additional sample
 collection.    A  chain of  custody form  should
 accompany  each  cooler  of samples  and  should
 include the  trip standard that is  dedicated to that
 group of samples.

 3.9.7  Tedlar  Bag Check

 Prior to use, one bag should be removed from each
 lot (case of 100) of Tedlar bags to be used for
 sampling and checked for possible contamination as
 follows: the test bag should be filled with ultra-zero
 grade air; a sample should be drawn from the bag
 and analyzed via Photovac GC or  whatever method
 is to be used for sample analysis. This procedure
 will ensure sample container cleanliness prior to the
 start of the sampling effort.

 3.9.8  SUMMA Canister Check

 From each lot of  four cleaned SUMMA canisters,
 one is to be removed for a GC/MS certification
 check.  If the canister passes certification, then it is
 re-evacuated and all four canisters from that lot are
 available for sampling.

If the  chosen canister is contaminated, then  the
 entire  lot of four SUMMA canisters  must be
recleaned, and a  single canister  is  re-analyzed by
GC/MS for certification.

3.9.9 Options

Duplicate Samples

A minimum of  5%  of all  samples should  be
collected in duplicate (i.e., if a total of 100 samples
are to  be  collected,  five  samples  should  be
duplicated).    In  choosing  which  samples  to
duplicate, the following criterion  applies:  if, after
filling the first Tedlar bag, and, evacuating the well
for 15  seconds, the second HNU (or other field
monitoring device being used) reading matches or
is close  to (within 50%)  the  first  reading, a
duplicate sample may be taken.
Spikes

A Tedlar bag spike and Tenax tube spike may be
desirable in situations where high concentrations of
contaminants other than the target compounds are
found to exist (landfills, etc.).  The additional level
of QA/QC attained by this practice can be useful in
determining the effects of interferences caused by
these non-target compounds.   SUMMA canisters
containing samples are not spiked.
3.10   DATA VALIDATION

For   each  target  compound,  the   level  of
concentration found hi the sample must be greater
than three times the level (for that  compound)
found in the field blank which accompanied that
sample to be considered valid.  The same criteria
apply to target compounds detected in the Tedlar
bag pre-sampling contamination check.
3.11   HEALTH AND SAFETY

Because   the   sample  is   being   drawn  from
underground, and no contamination is introduced
into the breathing  zone, soil gas sampling usually
occurs in Level D, unless the sampling location is
within the hot  zone of a site, which requires Level
B or Level C protection.  However, to ensure that
the proper level of protection is utilized, constantly
monitor the ambient air using the HNU PI 101 to
obtain background readings during the  sampling
procedure. As long as the levels in  ambient air do
not rise above background, no upgrade of the level
of protection is needed.

Also, perform an underground utility search prior to
sampling (see section 3.4.4).  When working with
potentially hazardous materials,  follow U.S. EPA,
OSHA, and specific health and safety procedures.
                                                17

-------


-------
                   4.0   General  Surface Geophysics:  SOP #2159
 4.1     SCOPE AND APPLICATION

 This Standard Operating Procedure (SOP) describes
 the general procedures used  to acquire  surface
 geophysical data.  This data is used for delineating
 subsurface  waste,  and for interpreting geologic,
 hydrogeologic or other data related to hazardous
 waste site characterization.

 The media pertinent to these surface geophysical
 methods are  soil/rock and groundwater.  The
 sensitivity or minimum response of a given method
 depends on the comparison of the object or area of
 study  to that  of its background (i.e., what the
 media's response would be like without the object
 of study).  Therefore, the suitability of  surface
 geophysical methods for a given investigation must
 be judged on the object's ability to be measured and
 the extent to which the specific setting of the study
 interferes with the measurement.

 The surface geophysical  method(s) selected for
 application  at  a  site are  dependent on site
 conditions,  such  as depth to bedrock, depth  to
 target,  urban  disturbances  (fences, power lines,
 surface debris, etc.) and  atmospheric conditions.
 Detectability of the target  is  dependent  on the
 sensitivity of the instrument and the variation of the
 field   measurement  from  the  ambient   noise.
 Ambient noise is the pervasive noise associated with
 an environment.   Therefore, the applicability  of
 geophysical methods at a given site is dependent on
 the specific setting at that site.

 Five  geophysical  methods  may be  utilized  in
 hazardous    waste    site   characterization:
 magnetometry,  electromagnetics,   resistivity,
 seismology and ground penetrating radar  (GPR).
 Magnetometers may  be  used  to  locate  buried
 ferrous metallic objects and geologic information.
 Electromagnetic methods can be used to determine
 the presence of metals, electrical conductivity of the
 terrain,  and geologic  information.   Resistivity
 methods  are  used to  determine  the electrical
 resistivity of the terrain and geologic information.
 Seismic methods are useful in determining geologic
 stratigraphy  and structure.  GPR may be used to
 locate disturbance in the soil (i.e., trenches,  buried
utilities and fill boundaries) and some near-surface
geologic information.
 These procedures may be varied  or changed as
 required, dependent on site conditions, equipment
 limitations or limitations imposed by the procedure.
 In all instances, the procedures employed should be
 documented and associated with the final report.
 4.2    METHOD SUMMARY

 4.2.1   Magnetics

 A magnetometer is an instrument which measures
 magnetic  field  strength  in  units  of  gammas
 (nanoteslas).  Local variations, or anomalies, in the
 earth's magnetic field are the result of disturbances
 caused mostly by variations in concentrations of
 ferromagnetic material  in  the  vicinity  of  the
 magnetometer's sensor.  A buried ferrous object,
 such as a steel drum or tank, locally distorts the
 earth's magnetic field and results in a magnetic
 anomaly.  The objective  of conducting a magnetic
 survey  at  a  hazardous waste  or  groundwater
 pollution site is  to  map  these  anomalies  and
 delineate the area containing buried sources of the
 anomalies.

 Analysis of magnetic data can allow an experienced
 geophysicist to estimate the areal extent of buried
 ferrous targets, such as a steel  tank  or  drum.
 Often,  areas  of burial  can be prioritized upon
 examination of the data, with high priority areas
 indicating  a  near certainty  of  buried ferrous
 material.  In some instances, estimates of depth of
 burial can be  made from the data.  Most of these
 depth  estimates  are   graphical   methods   of
 interpretation, such as slope techniques  and half-
 width rules, as described  by Nettleton (1976). The
 accuracy of these methods is dependent upon the
 quality of the  data and the skill of the interpreting
 geophysicist.   An accuracy of 10 to 20  percent is
 considered acceptable. The magnetic method may
 also be used to map certain geologic features, such
 as igneous intrusions, which may play an important
 role in the hydrogeology of a groundwater pollution
 site.

Advantages

Advantages of using the  magnetic method for  the
initial assessment of hazardous waste sites are  the
                                                 19

-------
relatively low cost of conducting the survey and the
relative ease of completing  a survey in a short
amount of time.  Little, if any, site preparation is
necessary.   Surveying requirements are  not  as
stringent  as for  other   methods  and  may  be
completed with a transit  or  Brunton-type pocket
transit and a non-metallic measuring tape.  Often,
a magnetic investigation  is  a very cost-effective
method for initial assessment of a hazardous waste
site  where  buried steel  drums  or tanks  are a
concern.

D/sac/vanfages

"Cultural  noise" is a limitation of the  magnetic
method in certain areas. Man-made structures that
are constructed with ferrous material, such as steel,
have a detrimental effect on the quality of the data.
Avoid features such as steel structures, power lines,
metal fences, steel reinforced concrete,  pipelines
and underground utilities.  When these features are
unavoidable, note their locations in a field notebook
and on the site map.

Another limitation of the magnetic method is the
inability   of   the   interpretation   methods  to
differentiate between various steel objects.   For
instance,  it is  not  possible  to determine if  an
anomaly is the  result of a steel tank, or a group of
steel drums, or old washing  machines.  Also, the
magnetic method does not allow the interpreter to
determine the contents of a buried tank or drum.

4.2.2  Electromagnetics

The  electromagnetic  method  is  a  geophysical
technique  based  on the physical  principles  of
inducing and detecting electrical current flow within
geologic strata.  A receiver detects these induced
currents by measuring the resulting time-varying
magnetic field.   The  electromagnetic method
measures  bulk   conductivity  (the   inverse  of
resistivity)  of  geologic   materials  beneath  the
transmitter and receiver  coils.  Electromagnetics
should not be confused with the electrical resistivity
method. The difference between the two techniques
is in the method which the electrical currents are
forced to flow in the earth. In the electromagnetic
method, currents are induced by the application of
time-varying magnetic  fields,  whereas  in  the
electrical resistivity method, current is injected into
the ground through surface electrodes.

Electromagnetics can be used to locate pipes, utility
lines, cables, buried steel drums, trenches, buried
waste, and concentrated contaminant plumes. The
method can also be used to map shallow geologic
features, such as lithologic changes and fault zones.

Advantages

Electromagnetic measurements can be collected
rapidly  and  with  a  minimum number of field
personnel. Most electromagnetic equipment used in
groundwater pollution investigations is lightweight
and easily portable. The electromagnetic method is
one of  the more  commonly  used  geophysical
techniques  applied  to  groundwater  pollution
investigations.

Disadvantages

The main limitation of the electromagnetic method
is "cultural noise".   Sources of "cultural noise" can
include:  large metal objects, buried cables, pipes,
buildings, and metal fences.

The electromagnetic method has limitations in areas
where the geology  varies laterally. These can cause
conductivity anomalies or lineations, which might be
misinterpreted as contaminant plumes.

4.2.3  Electrical  Resistivity'

The electrical resistivity method is; used to map
subsurface electrical resistivity structure, which is in
turn interpreted by the geophysicist to determine
the geologic structure and/or physical properties of
the  geologic materials.  Electrical 'resistivities of
geologic materials are measured in ohm-meters, and
are  functions  of  porosity, permeability,  water
saturation and the  concentration of dissolved solids
in the pore fluids.

Resistivity methods measure the bulk resistivity of
the subsurface, as  do the electromagnetic methods.
The difference  between the two methods is in the
way that electrical  currents are forced to flow in the
earth. In the electrical resistivity method, current is
injected into the ground through surface electrodes,
whereas hi  electromagnetic methods currents are
induced  by application  of  time-varying magnetic
fields.

Advantages

The principal advantage  of the electrical resistivity
method  is that quantitative modeling  is possible
                                                  20

-------
 using either computer software or published master
 curves. The resulting models can provide accurate
 estimates of depths, thicknesses and resistivities of
 subsurface layers. The layer resistivities can then be
 used to estimate the resistivity of the saturating
 fluid, which is related to the total concentration of
 dissolved solids in the fluid.

 Disadvantages

 The limitations of using the  resistivity method in
 groundwater pollution site investigations are largely
 due to  site characteristics,  rather than  in any
 inherent  limitations  of the  method.   Typically,
 polluted sites are located  in  industrial areas that
 contain an abundance of broad spectrum electrical
 noise.   In  conducting a  resistivity  survey, the
 voltages are relayed to the  receiver over long wires
 that are grounded at each end. These  wires act as
 antennae receiving the radiated electrical noise that
 in  turn degrades  the  quality  of  the measured
 voltages.

 Resistivity surveys require a fairly large area, far
 removed  from pipelines and grounded  metallic
 structures  such  as  metal  fences,  pipelines  and
 railroad tracks.  This requirement precludes  using
 resistivity on many polluted sites.  However, the
 resistivity method can often be used successfully off-
 site to map the stratigraphy of the area surrounding
 the site. A general "rule of thumb" for resistivity
 surveying is that grounded structures be at least half
 of the maximum electrode spacing distance  away
 from the axis of the survey line.

Another consideration in the  resistivity method is
 that the fieldwork tends to be  more labor intensive
 than  some  other geophysical techniques.    A
 minimum  of two  to three  crew  members are
 required for the fieldwork.

 4.2.4  Seismic

Surface seismic techniques used in groundwater
pollution site investigations are largely restricted to
seismic refraction and seismic reflection methods.
The  equipment   used  for  both  methods  is
fundamentally the same and both methods measure
the  travel-time  of  acoustic  waves  propagating
through the subsurface. In the refraction method,
the travel-time of waves refracted along an acoustic
interface is measured, and in the reflection method,
the travel-time of a wave which reflects or  echoes
off an interface is measured.
 The  interpretation  of  seismic  data will yield
 subsurface velocity information, which is dependent
 upon  the acoustic properties of the subsurface
 material.   Various geologic materials can be
 categorized by their acoustic properties or velocities.
 Depth to geologic interfaces are calculated using the
 velocities obtained from a  seismic investigation.
 The geologic information gained from  a seismic
 investigation  is then used in the  hydrogeologic
 assessment of a groundwater pollution site and the
 surrounding  area.   The interpretation of seismic
 data  indicates changes in lithology or stratigraphy,
 geologic structure, or water saturation (water table).
 Seismic methods are commonly used to determine
 the   depth   and   structure   of  geologic  and
 hydrogeologic  units,   to   estimate   hydraulic
 conductivity, to detect cavities or voids, to determine
 structure stability, to detect fractures  and fault
 zones, and to estimate ripability.  The  choice of
 method depends upon the information needed and
 the nature of the study area. This decision must be
 made by a geophysicist who is experienced in both
 methods, is  aware of the geologic  information
 needed by the hydrogeologist, and is also aware of
 the environment of the study area. The refraction
 technique  has been used more often  than  the
 reflection  technique  for  hazardous  waste  site
 investigations.

 Seismic Refraction Method

 Seismic refraction is most commonly used at sites
 where bedrock is  less than  500 feet below the
 ground surface.  Seismic refraction is simply the
 travel  path of a  sound wave through an  upper
 medium and along an interface and then back to the
 surface.   A  detailed discussion of the seismic
 refraction technique can be found in Dobrin (1976),
 Telford, et. al. (1985), and Musgrave (1967).

Advantages: Seismic refraction surveys are more
 common   than   reflection   surveys    for   site
 investigations.  The velocities of each layer can be
 determined from refraction data, and  a relatively
 precise estimate of the depth to different interfaces
 can be calculated.

 Refraction surveys add to depth information  in-
 between boreholes. Subsurface information can be
 obtained between boreholes at a fraction of the cost
 of drilling.    Refraction  data can be  used  to
 determine the depth to the water  table  or bedrock.
 In buried valley areas, refraction surveys map the
 depth  to bedrock.   The velocity  information
                                                  21

-------
obtained from a refraction survey can be related to
various physical properties of the bedrock.  Rock
types have certain ranges of velocities  and these
velocities are not always unique to a particular rock
type.   However, they can allow a geophysicist to
differentiate between certain units, such as shales
and granites.

Disadvantages:  The seismic refraction method
is based  on several assumptions.  To successfully
resolve the subsurface using the refraction method,
the conditions  of the geologic  environment must
approximate these assumptions:

    •   the velocities of the layers increase with
        depth,

    •   the velocity  contrast between layers is
        sufficient to resolve the interface, and

    •   the geometry of the geophones in relation
        to the refracting layers will permit the
        detection of thin layers.

These conditions must be met for accurate depth
information.

Collecting and  interpreting seismic refraction data
has several disadvantages.  Data collection can be
labor intensive. Also, large line lengths are needed;
therefore, as a  general rule, the distance from the
shot, or seismic source, to the first geophone station
must be  at least three times the desired depth of
exploration.

Seismic Reflection Method

The seismic reflection method is not as  commonly
used on groundwater pollution site investigations as
seismic refraction. In the seismic reflection method,
a sound  wave travels  down to a geologic interface
and reflects back to the surface.  Reflections occur
at an interface where there is a change in the
acoustic  properties of the subsurface material.

Advantages:   The seismic  reflection  method
yields information that allows  the interpreter to
discern between fairly discrete layers, so it is useful
for mapping stratigraphy. Reflection data is usually
presented in profile form, and depths to interfaces
are represented as  a function  of time.   Depth
information can be obtained by converting time
sections  into depth measurements using velocities
obtained from seismic refraction data, sonic logs, or
velocity logs.  The reflection technique requires
much less space than refraction surveys. The long
offsets of the seismic source from the geophones,
common hi refraction surveys, are not required in
the  reflection  method.     In  some  geologic
environments, reflection data can yield acceptable
depth estimates.

Disadvantages:   The major  disadvantage  to
using  reflection  data  is that a precise  depth
determination cannot be made. Velocities obtained
from most reflection data are at least 10% and can
be 20% of the true velocities. The interpretation of
reflection data requires a qualitative approach.  In
addition  to  being  more  labor intensive,  the
acquisition of reflection data is more complex than
refraction data.

The reflection method places higher requirements
on  the  capabilities of  the  seismic equipment.
Reflection data is commonly used  in the petroleum
exploration industry and requires a large amount of
data processing  time  and lengthy data collection
procedures.   Although mainframe computers are
often used in the reduction and  analysis of large
amounts of reflection data,  recent advances have
allowed for the use of personal computers on small
reflection surveys for engineering purposes. In most
cases,  the data  must be recorded  digitally  or
converted to a digital  format,  to employ various
numerical processing operations.  The use of high
resolution reflection seismic method? relies heavily
on the geophysicist,  the computer capacity, the data
reduction and processing  programs,  resolution
capabilities of the seismograph and geophones, and
the  ingenuity  of  the interpreter.  Without these
capabilities,   reflection  surveys    are   not
recommended.

4.2.5  Ground Penetrating Radar

The ground penetrating  radar (GPR) method is
used for a variety of civil  engineering, groundwater
evaluation and hazardous waste  site applications.
This geophysical method is the most site-specific of
all  geophysical  techniques,  providing subsurface
information ranging in depth from several tens of
meters to only a fraction of a  meter.  A basic
understanding  of  the  function of  the  GPR
instrument,  together  with  a knowledge of the
geology  and  mineralogy of  the site, can  help
determine if GPR  will be  successful  in  the site
assessment.   When possible, the GPR technique
should be integrated with other  geophysical and
                                                  22

-------
 geologic data to provide the most comprehensive
 site assessment.

 The GPR method uses a transmitter that emits
 pulses of high-frequency electromagnetic waves into
 the subsurface.  The transmitter is either moved
 slowly across the ground surface or moved at fixed
 station intervals. The penetrating electromagnetic
 waves are scattered  at points of  change in the
 complex dielectric permittivity, which is a property
 of the subsurface material dependent primarily upon
 the bulk density, clay content and water content of
 the   subsurface   (Olhoeft,    1984).      The
 electromagnetic energy which is scattered back to
 the receiving antenna on the surface is recorded as
 a function of tune.

 Depth penetration is severely limited by attenuation
 of the transmitted electromagnetic waves  into the
 ground.   Attenuation is  caused by the  sum  of
 electrical conductivity,  dielectric  relaxation, and
 geometric scattering losses in  the  subsurface.
 Generally, penetration of  radar  frequencies  is
 minimized by a shallow water table, an increase in
 the  clay content  of  the  subsurface,   and  in
 environments where the electrical resistivity of the
 subsurface is less than 30 ohm-meters (Olhoeft,
 1986). Ground penetrating radar works best in dry
 sandy  soil above the water table.  At applicable
 sites, depth resolution should be between 1 and 10
 meters (Benson, 1982).

 The  analog  plot  produced by  a continuously
 recording GPR system is analogous to a seismic
 reflection profile; that is, data is represented as a
 function  of horizontal distance  versus time.  This
 representation  should not  be confused  with a
 geologic  cross section which  represents data as a
 function   of  horizontal  distance  versus  depth.
 Because very high-frequency electromagnetic waves
 in the megahertz range are used by radar systems,
 and time delays are measured in nanoseconds (10~9
 seconds), very high resolution of the subsurface is
 possible using GPR. This resolution can be as high
 as  0.1 meter.   For  depth determinations,  it  is
 necessary to correlate the recorded features  with
 actual depth measurements from boreholes or from
 the results  of  other  geophysical  investigations.
When properly interpreted, GPR data can optimally
resolve changes in  soil horizons,  fractures, water
insoluble  contaminants,  geological features, man-
made buried objects, and hydrologic features  such
as water  table depth and wetting fronts.
Advantages

Most GPR systems can provide a continuous display
of  data along  a traverse  which  can often  be
interpreted qualitatively in  the  field.   GPR is
capable  of providing high resolution data  under
favorable site conditions.  The real-time capability
of GPR results in a rapid turnaround, and allows
the geophysicist to quickly evaluate subsurface site
conditions.

Disadvantages

One of the major limitations of GPR is the site-
specific nature of the technique. Another limitation
is the cost of site preparation which is necessary
prior to the survey.   Most GPR units are towed
across the ground  surface.  Ideally, the ground
surface should be flat, dry, and clear of any brush or
debris.  The quality of the data can be degraded by
a variety of factors, such as an  uneven ground
surface or various cultural noise sources. For these
reasons, it  is mandatory that  the site be visited by
the project geophysicist before a GPR investigation
is proposed.  The geophysicist should also evaluate
all  stratigraphic  information available,  such  as
borehole data and information on the depth to
water table in the survey area.
4.3    SAMPLE  PRESERVATION,
        CONTAINERS, HANDLING AND
        STORAGE

This section is not applicable to this SOP.
4.4    INTERFERENCES AND
        POTENTIAL PROBLEMS

See section 4.2.1 for a discussion  of limitations of
the magnetic method.

See section 4.2.2 for a discussion  of limitations of
the electromagnetic method.

See section 4.2.3 for a discussion  of limitations of
the electrical resistivity method.

See section 4.2.4 for a discussion  of limitations of
the seismic refraction method and  the  seismic
reflection method.
                                                 23

-------
See section 4.2.5 for a discussion of limitations of
the ground penetrating radar (GPR) method.
4.5    EQUIPMENT/APPARATUS

4.5.1  Magnetics

    •  GEM GSM-19G
       magnetometer/gradiometer, EDA OMNI
       IV magnetometer/gradiometer,
       Geonics 856AGX (with built-in datalogger)
       or equivalent
    •  magnetometer base station
    •  300-foot tape measure
    •  non-ferrous  survey  stakes (wooden or
       plastic)

4.5.2  Electromagnetics
       Geonics EM-31, EM-34 or equivalent
       Polycorder datalogger
       Dat 31Q software (data dump software)
       300-foot tape measure
       survey stakes
4.5.3 Electrical Resistivity

    •  DC resistivity unit (non-specific)
    •  4 electrodes and appropriate cables (length
       dependent on depth of survey)
    •  1 or 2 12-volt car batteries
    •  300-foot tape measure

4.5.4 Seismic
        12- or 24-channel seismograph (Geometries
        2401 or equivalent)
        30    lOHz  to  MHz geophones  (for
        refraction)
        30    50Hz  or greater geophones  (for
        reflection)
        300-foot tape measure
        survey stakes
        sledge  hammer  and  metal   plate  or
        explosives
4.5.5  Ground Penetrating Radar
        GSSI SIR-8 or equivalent
        80   Mhz,   100  Mhz  or
        antenna/receiver pit
        200-foot cable
        300-foot tape measure
300  Mhz
                4.6    REAGENTS

                This section is not applicable to this SOP.


                4.7    PROCEDURES

                Refer to the manufacturer's operating manual for
                specific procedures  relating to operation of the
                equipment.


                4.8    CALCULATIONS

                Calculations vary based on the geophysical method
                employed.  Refer to the instrument-specific users
                manual for specific formulae.


                4.9    QUALITY ASSURANCE/
                       QUALITY CONTROL

                The following general quality  assurance  activities
                apply to the implementation of these procedures.

                    •  All data must be documented on field data
                       sheets or within site logbooks.

                    •  All instrumentation must be operated in
                       accordance  with operating instructions as
                       supplied  by  the  manufacturer,  unless
                       otherwise specified in  the work  plan.
                       Equipment   checkout   and  calibration
                       activities  must  occur  prior  to
                       sampling/operation, and they  must  be
                       documented.             ,

                Method-specific quality assurance procedures may
                be found in the user's manual.
                4.10  DATA VALIDATION

                Evaluate  data  as per the criteria established  in
                section 4.9 above.
4.11   HEALTH AND SAFETY

When working with potentially hazardous materials,
follow U.S. EPA, OSHA and specific health and
safety procedures.
                                              24

-------
APPENDIX A




   Figures
     25

-------
    Figure 1:  Sampling Augers

          SOP #2012
 TUBE
<;UGER

BUCKE
 AUGER
             26

-------
    Figure 2:  Sampling Trier


         SOP #2012
     d
IT
 i V
5 .Si
           -j  L
                  1.27-2.54 cm
              27

-------
                             Figure 3: Sampling Train Schematic

                                        SOP #2149
VACUUM
  BO
                 SAMPLING
                   PORT
                                                           1/4" TEFLON TUBING
                                             1/4"  I.D. THIN WALL
                                               TEFLON  TUBING
                                                   1/4" S.S.
                                                SAMPLE PROBE
"QUICK CONNECT'
    FITTING
                                                                             MODELING
                                                                               CLAY
                                                          SAMPLE
                                                           WELL
                                            28

-------
 APPENDIX B



HNU Field Protocol
       29

-------
                                         HNU Field Protocol
                                            SOP  #2149
 Startup Procedure

 1.  Before attaching the probe, check the function
     switch on the control panel to ensure that it is
     in the "off position.   Attach the probe by
     plugging it into the interface on the top of the
     readout  module.   Use care in aligning the
     prongs in the probe cord with the socket: do
     not force it.

 2.  Turn the function  switch to the battery check
     position.  The needle on the meter should read
     within or above the green area on the scale. If
     not, recharge the battery.  If the red indicator
     light comes on,  the battery needs recharging.

 3.  Turn the function switch to any range setting.
     For no more than 2 to 3 seconds, look into the
     end of the probe to see if the  lamp is on.  If it
     is on, you will see a purple glow.  Do not stare
     into the probe any longer  than three seconds.
     Long term exposure to UV light can damage
     the eyes.  Also, listen  for  the hum of the fan
     motor.

 4.   To zero the instrument, turn the function switch
     to the standby  position and  rotate the zero
     adjustment until the meter  reads zero.   A
     calibration gas is not needed  since this is an
     electronic zero  adjustment.    If  the  span
     adjustment setting is changed after the zero is
     set, the zero should  be rechecked and adjusted,
     if necessary.  Wait  15 to 20 seconds to ensure
     that the zero reading is stable.  If necessary,
    readjust the zero.

 Operational Check

 1.   Follow the startup procedure.

2.   With the instrument set on the 0-20 range, hold
    a solvent-based Magic Marker near the probe
    tip.    If  the meter  deflects  upscale,  the
    instrument is working.

Field Calibration Procedure

1.   Follow  the  startup  procedure   and   the
    operational check.
 2.  Set the function switch to the range setting for
     the concentration of the calibration gas.

 3.  Attach  a  regulator  (HNU  101-351)  to a
     disposable cylinder of isobutylene gas. Connect
     the regulator to the probe of the HNU with a
     piece of clean Tygon tubing.  Turn the valve on
     the regulator to the "on" position.

 4.  After 15 seconds, adjust the span dial until the
     meter reading equals the concentration of the
     calibration gas used.   The calibration gas is
     usually 100 ppm of isobutylene in zero air.  The
     cylinders are marked in benzene equivalents for
     the 10.2 eV probe  (approximately  55 ppm
     benzene equivalent) and for the 11.7 eV probe
     (approximately 65  ppm benzene  equivalent).
     Be  careful to  unlock  the span  dial before
     adjusting it. If the span has to be set below 3.0
     calibration, the lamp and ion chamber should
     be inspected and cleaned as appropriate.  For
     cleaning of the 11.7 eV  probe, only  use an
     electronic-grade, oil-free freon or similar water-
     free, grease-free solvent.

 5.   Record in the field log:  the  instrument ID #
     (EPA decal or serial number if the instrument
     is a rental); the initial and final span settings;
     the date and  time; concentration and type of
     calibration used; and the name of the person
     who calibrated the instrument.

 Operation

 1.   Follow  the  startup  procedure,   operational
     check, and calibration check.

 2.   Set  the  function switch to  the  appropriate
     range. If the concentration of gases or vapors
     is unknown, set the function switch to the 0-20
     ppm range.  Adjust it as necessary.

3.    While taking care not to permit the HNU to be
     exposed  to   excessive   moisture,  dirt,  or
     contamination, monitor the work  activity as
     specified in the site health and safety plan.

4.   When the activity is completed or at the end of
    the day, carefully clean the outside of the HNU
    with a damp disposable towel to remove any
                                                 30

-------
    visible dirt.  Return the HNU to a secure area       plastic to prevent it from becoming contaminated
    and place on charge.                               and to prevent water  from getting inside in the
                                                       event of precipitation.
5.  With  the exception  of the probe's inlet and
    exhaust, the HNU can be wrapped in clear
                                                   31

-------

-------
                                        References


SOPS #2006, 2012, 2149

American Standards for Testing and Materials.  1988. Standard Method for Preparing Test and Split-
        Barrel Sampling of Soils:  Annual Book of ASTM Standards. Section 4, Volume 4.08. ASTM
        D1586-84.

Earth, D.S. and BJ. Mason. 1984. Soil Sampling Quality Assurance User's Guide.  EPA/600/4-
        84/043.

de Vera, E.R., B.P. Simmons, R.D. Stephen, and D.L. Storm. 1980. Samplers and Sampling
        Procedures for Hazardous Waste Streams.  EPA/600/2-80/018.

Gilian Instrument Corp. 1983.  Instruction Manual for Hi Flow Sampler: HFS 113, HFS 113 T, HFS
        113 U, HFS 113 UT.

HNU Systems, Inc. 1975. Instruction Manual for Model PI 101 Photoionization Analyzer.

Mason, BJ. 1983. Preparation of Soil Sampling Protocol: Technique and Strategies. EPA/600/4-
        83/020.

National Institute for Occupational Safety and Health. October, 1985.  Occupational Safety and Health
        Guidance Manual for Hazardous Waste Site Activities. NIOSH/OSHA/USCG/EPA.

New Jersey Department of Environmental Protection.  February, 1988.  Field Sampling Procedures
        Manual.

Roy F. Weston, Inc. 1987. Weston Instrumentation Manual, Volume I.

U.S. Environmental Protection Agency. December, 1984. Characterization of Hazardous Waste Sites -
        A Methods Manual: Volume II, Available Sampling Methods,  2nd Edition.  EPA/600/4-
        84/076.

U.S. Environmental Protection Agency. April 1, 1986. Engineering Support Branch Standard
        Operating Procedures and Quality Assurance Manual.  U.S. EPA Region IV.

U.S. Environmental Protection Agency. 1987. A Compendium of Superfund Field Operations
        Methods.  EPA/540/P-87/001. Office  of Emergency and Remedial Response. Washington,
        D.C. 20460.
 SOP #2159

        Magnetics

 Breiner, S. 1973. Applications Manual for Portable Magnetometers:  EG&G GeoMetrics. Sunnyvale,
        California.

 Fowler, J. and D. Pasicznyk.  February, 1985.  Magnetic Survey Methods Used in the Initial Assessment
        of a Waste Disposal Site:  National Water Well Association Conference on Surface and
        Borehole Geophysics.
                                              33

-------
 Lilley, F. 1968. Optimum Direction of Survey Lines.  Geophysics 33(2): 329-336.

 Nettleton, L.L. 1976.  Elementary Gravity and Magnetics for Geologists and Seismologists:  Society of
         Exploration Geophysicists.  Monograph Series Number L

 Redford, M.S. 1964.  Magnetic Anomalies over Thin Sheets. Geophysics 29(4): 532-536.

 Redford, M.S. 1964.  Airborne Magnetometer Surveys for Petroleum. Exploration: Aero Service
         Corporation.  Houston, Texas.

 Vacquier, V. and others.  195L  Interpretation of Aeromagnetic Maps: Geological Society of America.
         Memoir Number 47.


         Electromagnetics

 Duran, P.B.  1982.  The Use of Electromagnetic Conductivity Techniques in the Delineation of
         Groundwater Pollution Plumes: unpublished master's thesis, Boston University.

 Grant, F.S. and G.F. West. 1965. Interpretation Theory in Applied Geophysics.  McGraw-Hill Book
         Company, New York, New York.

 Greenhouse, J.P., and D.D. Slaine. 1983. The Use of  Reconnaissance Electromagnetic Methods to
         Map Contaminant Migration.  Ground Water Monitoring Review 3(2).                  \

 Keller, G.V. and F.C. Frischknecht. 1966. Electrical Methods in Geophysical Prospecting.  Pergamon
         Press, Long Island City, New York.

 McNeill, J.D. 1980. Electromagnetic Terrain Conductivity Measurements at Low Induction Numbers.
        Technical Note TN-6, Geonics  Limited. Mississauga, Ontario, Canada.

 McNeill, J.D. 1980. EM34-3 Survey Interpretation Techniques. Technical Note TN-8, Geonics
        Limited. Mississauga, Ontario, Canada.

 McNeill, J.D. 1980. Electrical Conductivity of Soils and Rocks. Technical Note TN-5, Geonics
        Limited. Mississauga, Ontario, Canada.

 McNeill, J.D. and M. Bosnar. 1986. Surface and Borehole Electro-Magnetic Groundwater
        Contamination Surveys, Pittman Lateral Transect, Nevada:  Technical Note TN-22, Geonics
        Limited. Mississauga, Ontario, Canada.

 Stewart, M.T. 1982. Evaluation of Electromagnetic Methods for Rapid Mapping of Salt Water
        Interfaces hi Coastal Aquifers.  Ground Water 20.

 Telford, W.M.,  L.P. Geldart, R.E. Sheriff, and DA. Keys.  1977. Applied Geophysics.  Cambridge
        University Press.  New York, New York.


        Electrical Resistivity

Bisdorf, RJ.  1985.  Electrical Techniques for Engineering Applications. Bulletin of the Association of
       Engineering Geologists 22(4).
                                              34

-------
Grant, F.S. and G.F. West.  1965. Interpretation Theory in Applied Geophysics.  McGraw-Hill Book
        Company, New York, New York.

Keller, G.V. and EC. Frischnecht.  1966. Electrical Methods in Geophysical Prospecting.  Pergamon
        Press, Long Island City, New York.

Kelly, W.E. and R.K. Frohlich. 1985. Relations between Aquifer Electrical and Hydraulic Properties.
        Ground Water 23:2.

Stollar, R. and P. Roux.  1975. Earth Resistivity Surveys - A Method for Defining ^Groundwater
        Contamination.  Ground Water 13.

Sumner, J.S.  1976. Principles of Induced Polarization for Geophysical Exploration. Elsevier Scientific
        Publishing, New York, New York.

Telford, W.M., L.P. Geldart, R.E. Sheriff, and DA. Keys. 1977. Applied Geophysics. Cambridge
        University Press, New York, New York.

Urish, D.W.  1983. The Practical Application of Surface Electrical Resistivity to Detection of Ground
        Water Pollution.  Ground Water 21.

Van Nostrand, R.E., and L.K. Cook.  1966.  Interpretation of Resistivity Data:  U.S. Geological Survey
        Professional Paper 499, Washington, D.C.

Zohdy, AA.R.  1975.  Automatic Interpretation of Schlumberger Sounding Curves Uusing Modified
        Dar Zarrouk Functions.  U.S. Geological Survey Bulletin 1313-E, Denver,  Colorado.
        Seismic

Coffeen, JA. 1978. Seismic Exploration Fundamentals. PennWell Publishing, Tulsa, Oklahoma.

Dobrin, M.B.  1976. Introduction to Geophysical Prospecting; 3rd ed.  McGraw-Hill, New York, New
        York.

Griffiths, D.H. and R.E. King. 1981. Applied Geophysics for Geologists and Engineers. Second edition.
        Pergamon Press, Oxford, England.

Miller, R.D., S.E. Pullan, J.S. Waldner, and P.P. Haeni. 1986.  Field Comparison of Shallow Seismic
        Sources.  Geophysics 51(11): 2067-92.

Musgrave, A.W.  1967.  Seismic Refraction Prospecting. The Society of Exploration Geophysicists.
        Tulsa, Oklahoma.

Telford, W.M., L.P. Geldant, R.E. Sheriff, and DA. Keys.  1985. Applied Geophysics. Cambridge
        University Press, Cambridge, England.


        Ground Penetrating Radar

Benson, R.C., R A. Glaccum, and M.R. Noel. 1982.  Geophysical Techniques for Sensing Buried Wastes
        and Waste Migrations. Technos Inc.  Miami, Florida.  236 pp.
                                               35

-------
Olehoft, G.R.  1984. Applications and Limitations of Ground Penetrating Radar: Expanded Abstracts,
        Society of Exploration Geophysicists. 54th Annual Meeting: December 2-6, 1984.  Atlanta,
        Georgia. 147-148.
•U.S.Govarnment Printing Office: 1991 — 548-187/40577
                                                 36

-------