vvEPA  Method 1664:
       N-Hexane Extractable Material (HEM)
       and Silica Gel Treated N-Hexane
       Extractable Material (SGT-HEM)
       by Extraction and Gravimetry
       (Oil and Grease and Total Petroleum
       Hydrocarbons)
                                   i Printed on Recycled Paper

-------

-------
                                      Acknowledgments
This method was prepared under the direction of William A. Telliard of the Engineering and Analysis
  Division within the EPA Office of Water. This document was prepared under EPA Contract No.
                          68-C3-0337 by DynCorp Environmental, Inc.
                                         Disclaimer

   This method has been reviewed by the Engineering and Analysis Division, U.S. Environmental
 Protection Agency, and approved for publication.  Mention of trade names or commercial products
                   does not constitute endorsement or recommendation for use.

-------

-------
                                                                                    Method 1664
Introduction
As a party to the Montreal Protocol on Substances that Deplete the Ozone Layer and as required by
law under the Clean Air Act Amendments of 1990 (CAAA), the United States is committed to
controlling and eventually phasing out the use of chlorofluorocarbons (CFCs).  In support of these
efforts, Method 1664 was developed by the United States Environmental Protection Agency Office of
Science and Technology to  replace previously used gravimetric procedures that employed Freon-113, a
Class I CFC, as the extraction solvent for the determination of oil and grease and petroleum hydrocar-
bons.

Method 1664 is a performance-based method applicable to  aqueous matrices that requires the use of n-
hexane as the extraction solvent and gravimetry as the determinative  technique. Alternative extraction
and concentration techniques are allowed, provided that all  performance specifications are met.  In
addition, QC procedures designed to monitor precision and accuracy  have been incorporated into
Method 1664.

Questions concerning this method or its application should  be addressed to:

W.A. Telliard
USEPA Office of Water
Analytical Methods Staff
Mail Code 4303
401 M Street, SW
Washington,  D.C. 20460
202/260-7120
Requests for additional copies should be directed to:

Water Resource Center
Mail Code RC-4100
401 M Street,  SW
Washington, D.C. 20460
202/260-7786  or 202/260-2814

-------

-------
                                  Method  1664

     N-Hexane  Extractable  Material  (HEM) and Silica Gel Treated N-

 Hexane Extractable Material (SGT-HEM) by Extraction and Gravimetry

            (Oil and Grease and Total Petroleum Hydrocarbons)


1.0   Scope and Application

 1.1   This method is for determination of n-hexane extractable material (HEM) and n-hexane extracta-
      ble material that is not adsorbed by silica gel (SGT-HEM) in surface and saline waters and
      industrial and domestic aqueous wastes. Extractable materials that may be determined are
      relatively non-volatile hydrocarbons, vegetable oils, animal fats, waxes, soaps, greases, and
      related materials. The method is based on prior Environmental Protection Agency (EPA) and
      association methods for determination of "oil and grease"  and "total petroleum hydrocarbons"
      (References 16.1 and  16.2).
 1.2   This method is for use in the Environmental Protection Agency's (EPA's) survey and monitoring
      programs under the Clean Water Act.  "Oil and grease" is a conventional pollutant defined in the
      Act and codified at 40 CFR 401.16.  The term "n-hexane  extractable material" reflects that this
      method can be applied to materials other than oils and greases.  Similarly, the term "silica gel
      treated n-hexane extractable material"  reflects that this method can  be applied to materials other
      than aliphatic petroleum hydrocarbons that are not adsorbed by silica gel.
 1.3   This method is not applicable  to measurement of materials that volatilize at temperatures below
      approximately 85°C.  Petroleum fuels  from gasoline through #2 fuel oil may be  partially lost in
      the solvent removal operation.
 1.4   Some crude oils and heavy fuel oils contain a significant percentage of materials that are not
      soluble in n-hexane.  Accordingly, recoveries of these materials may be low.
 1.5  This method is capable of measuring HEM and SGT-HEM in the range of 5 to 1000 mg/L, and
     may  be extended to higher levels by analysis of a smaller  sample volume collected separately.
 1.6  For this method, the Method Detection Limit (MDL; 40 CFR 136, Appendix B)  has been
     determined as 1.4 mg/L for HEM and 1.6 mg/L for SGT-HEM (Reference 16.9), and the Mini-
     mum Level (ML; Reference 16.3) has been set at  5.0 mg/L for HEM and SGT-HEM (Reference
      16.9).
1.7  This  method is "performance-based".  The analyst is permitted to modify the method to over-
     come interferences or  lower the cost of measurements, provided that all performance criteria in
     this method are met.  The requirements for establishing method equivalency are  given in Section
     9.1.2.
1.8  Any modification of this method, beyond those expressly permitted, shall be considered a major
     modification subject to application and approval of alternate test procedures under 40 CFR 136.4
     and 136.5.

-------
  Method 1664
  1.9   Each laboratory that uses this method must demonstrate the ability to generate acceptable results
       using the procedure in Section 9.2.

 2.0   Summary of Method

  2.1   A 1-L sample is  acidified to pH <2 and serially extracted three times with n-hexane in a
       separatory funnel.  The extract is dried over sodium sulfate.
  2.2   The solvent is evaporated from  the extract and the HEM is weighed.  If the HEM is to be used
       for determination of SGT-HEM, the HEM is redissolved in n-hexane.
  2.3   For SGT-HEM determination, an amount of silica gel proportionate to the amount of HEM  is
       added to the solution containing the redissolved HEM to remove adsorbable materials. The
       solution is filtered to remove the silica gel, the solvent is evaporated, and the SGT-HEM is
       weighed.
 2.4   Quality is assured through calibration and testing of the extraction, concentration, and gravimet-
       ric systems.

3.0   Definitions

 3.1    HEM and SGT-HEM are method-defined analytes; i.e., the definitions of both HEM and SGT-
       HEM are dependent on the procedure used. The nature of the oils and/or greases, and the
       presence of extractable non-oily matter in the sample will influence the  material measured and
       interpretation of results.
 3.2    Definitions for terms used in this method are given in the glossary at the end of the method.

4.0    Interferences

 4.1   Solvents, reagents, glassware, and other sample-processing hardware may yield artifacts that
      affect results.  Specific selection of reagents and purification of solvents may be required.
 4.2   All materials used in the analysis shall be demonstrated to be free from  interferences under the
      conditions of analysis by running laboratory blanks as described  in Section 9.4.
 4.3   Glassware is cleaned by washing in hot water containing detergent, rinsing with tap and distilled
      water, and rinsing with solvent or baking.  Boiling flasks that will contain the extracted residue
      are dried in  an  oven at 105-115°C and stored in a desiccator.
 4.4   Sodium  sulfate and silica gel fines have the potential to inflate results for HEM and SGT-HEM
      by passing through the filter paper.  If the filter paper specified in this method is inadequate for
      removal of these fines, use of a 0.45-micron filter is recommended.
4.5   Interferences extracted from  samples will vary considerably from  source to source, depending
      upon the diversity of the site being sampled.  For those instances  in which samples are thought
      to consist of complex matrices containing substances  (such as particulates or detergents) that may
      interfere with the extraction procedure, a smaller sample may  need to be collected for analysis.

-------
                                                                                      Method 1664
5.0   Safety

 5.1   The toxicity or carcinogenicity of each reagent used in this method has not been precisely
      determined; however, each chemical should be treated as a potential health hazard.  Exposure to
      these chemicals should be reduced to the lowest possible level. It is suggested that the laborato-
      ry perform personal hygiene monitoring of each analyst using this method and that the results of
      this monitoring be made available to the analyst.
 5.2   n-Hexane has been shown to have increased neurotoxic  effects over other hexanes and some
      other solvents. Inhalation of n-hexane should be minimized by performing all operations with n-
      hexane in a hood or well-ventilated area.
 5.3   Unknown samples may contain high concentrations of volatile toxic compounds. Sample
      containers should be opened in a hood and handled with gloves to prevent exposure.
 5.4   This method does not address all safety issues associated with  its  use. The laboratory is respon-
      sible for maintaining a safe work environment and a current awareness file of OS HA regulations
      regarding the safe handling of the chemicals specified in this method. A reference file of
      material  safety data sheets (MSDSs) should be available to all  personnel involved in these
      analyses.  Additional information on laboratory safety can be found in References 16.4-16.6.

6.0   Equipment and Supplies

   Note:    Brand names, suppliers, and part numbers are for illustrative purposes only.  No
   endorsement is implied.  Equivalent performance may be achieved using apparatus and materials
   other than  those specified here, but demonstration of equivalent performance  that meets the
   requirements of this method is the responsibility of the laboratory.

 6.1   Sampling equipment.
       6.1.1  Sample collection bottles—Glass, approximately 1-L, with PTFE-lined screw cap.

   Note:    In those instances  necessitating collection of a smaller aliquot, a smaller sample
   container may be used.

       6.1.2  Cleaning.
            6.1.2.1  Bottles—Detergent water wash, tap water rinse, cap with aluminum foil, and bake
                    at 200-250°C for  1 h minimum prior to use. Solvent rinse may be used in place
                    of baking.
            6.1.2.2  Liners for screw caps—Detergent water wash, tap water and solvent  rinse, and
                    bake at 110-200°C for 1 h minimum prior  to use.
       6.1.3  Bottles and liners must be lot-certified to be free of artifacts by running laboratory  blanks
             according to this  method (per Section 9.4).  If blanks from bottles and/or liners without
             cleaning or with fewer cleaning steps  than required above  show no detectable materials,
             the bottle and  liner cleaning steps that do not  eliminate these artifacts  may be omitted.

-------
Method 1664
6.2   Equipment for glassware cleaning.
       6.2.1  Laboratory sink with overhead fume hood.
       6.2.2  Oven—Capable of maintaining a temperature within ± 5°C in the range of 100-250°C.
6.3   Equipment for calibration.
       6.3.1  Analytical Balance—Capable of weighing 0.1  mg.
       6.3.2  Volumetric flask—Glass,  100-mL.
       6.3.3  Vials—Assorted sizes, with PTFE-lined screw caps.
       6.3.4  Volumetric pipette—Glass, 5-rnL.
6.4   Equipment for sample extraction.
       6.4.1  Balance—Top loading, capable of weighing 500-2000 g within ±1%.
       6.4.2  Glass stirring rod.
       6.4.3  Separatory funnel—Glass, 2000-mL, with PTFE stopcock.
       6.4.4  Funnel—Large, glass, for  pouring sample into separatory funnel.
       6.4.5  Centrifuge—Capable of spinning at least four  100-mL glass centrifuge tubes  at 2400 rpm
             minimum.
       6.4.6  Centrifuge tubes—100-mL glass.
6.5   Equipment for removal of water, sodium sulfate, and silica gel fines.
       6.5.1  Funnel—Analytical, glass.
       6.5.2  Filter paper—Whatman No. 40 (or equivalent), to fit funnel.
6.6   Equipment for solvent evaporation.
       6.6.1  Water bath or Steam bath—Capable of maintaining a temperature of at least  85°C.
       6.6.2  Flask—Boiling,  125-mL (Corning No. 4100 or equivalent).
       6.6.3  Distilling head—Claisen (VWR Scientific No. 26339-005, or equivalent), includes
             Claisen-type connecting tube and condenser.
       6.6.4  Distilling adaptor (attached to  the distilling head and to the waste collection flask for
             recovery of solvent).
       6.6.5  Waste collection flask (attached to the distilling adaptor for collection of the  distilled
             solvent).
       6.6.6  Ice bath or recirculating chiller (to aid in the condensation and collection of the distilled
             solvent).
       6.6.7  Vacuum, pump or other source of vacuum.
       6.6.8  Tongs, for handling the boiling flask (Baxter Scientific Products No. T5007-2, or equiva-
             lent).
       6.6.9  Desiccator—Cabinet- or jar-type, capable of keeping the boiling flask (Section 6.6.2) dry
             during cooling.

-------
                                                                                      Method 1664
 6.7   Equipment for removal of adsorbable materials.
        6.7.1  Magnetic stirrer.
        6.7.2  PTFE-coated magnetic stirring bars.
        6.7.3  Graduated cylinder—500-mL, capable of measuring ± 5 mL.
        6.7.4  Pipettes—Assorted sizes, calibrated to within ± 0.5 percent.

 7.0   Reagents and Standards
 7.1   Reagent water—Water in which HEM is not detected at or above the Minimum Level of this
       method. Bottled distilled water, or water prepared by passage of tap water through activated
       carbon  have been shown to be acceptable sources of reagent water.
 7.2   Hydrochloric or sulfuric acid—ACS, 1:1. Mix equal volumes of concentrated HC1 or H2SO4 and
       reagent water.
 7.3   n-Hexane—85% purity, 99.0% min. saturated C6 isomers, residue less  than 1 mg/L.
 7.4   Acetone—ACS, residue less than 1 mg/L.
 7.5   Sodium sulfate—ACS, granular  anhydrous.

    Note:    Powdered sodium sulfate should not be used because traces of water may cause it to
    solidify.

 7.6   Boiling chips—Silicon carbide or fluoropolymer.
 7.7   Silica gel—Anhydrous, 75 - 150 micrometers, Davisil Grade 923 (Supelco 21447-7A, or equiva-
       lent). Dry at 200-250°C for 24  h  minimum and store in a desiccator or tightly sealed container.
       Determine the n-hexane soluble  material content of the silica gel by extracting 30 g of silica gel
       with n-hexane and evaporating to dryness.  The silica gel must contain less than 5 mg of n-
       hexane  soluble material per 30 g (< 0.17 mg/g).
 7.8   Hexadecane—98% minimum purity.
 7.9   Stearic  acid—98% minimum purity.
7.10   Hexadecane/stearic acid (1:1) spiking  solution—Prepare in acetone at a concentration of 4
       mg/mL each.
       7.10.1  Place 400 ± 4 mg stearic acid and 400 ± 4 mg hexadecane in a 100-mL volumetric flask
              and fill to the mark with acetone.

    Note:    The solution may require warming for complete dissolution of stearic acid.

       7.10.2  After the  hexadecane and stearic acid have dissolved, transfer the solution to a 100-150
              mL vial with fluoropolymer-lined cap. Mark the solution level on  the vial and store in
              the dark at room temperature.

-------
 Method 1664
       7.10.3  Immediately prior to use, verify the level on the vial and bring to volume with acetone,
              if required. Warm to redissolve all visible precipitate.

    Note:    If there is doubt of the concentration, remove 5.00 ± 0.05 mL with a volumetric pipet, place
    in a tared weighing pan, and evaporate to dryness in a fume hood.  The weight must be 40 ± 1 mg.

7.11   Precision and recovery (PAR) standard—Spike 5.00 ± 0.05 mL of the hexadecane/stearic acid
      spiking solution (Section 7.10) into 950-1050 mL of reagent water to produce concentrations of
      approximately 20 mg/L each of hexadecane and stearic acid.  The PAR standard is used for the
      determination of initial precision and recovery (Section 9.2.2) and ongoing precision and recovery
      (Section 9.6).
7.12   The spiking solutions should be checked frequently for signs of degradation or evaporation using
      the test noted in Section 7.10.3, and must be replaced after six months, or sooner if degradation has
      occurred.

8.0   Sample Collection, Preservation, and Storage

 8.1   Collect approximately one liter of representative sample in a glass bottle following conventional
      sampling practices (Reference 16.7), except that the bottle must not be pre-rinsed with sample
      before collection. To allow for potential QC failures, it is recommended that additional sample
      aliquots be collected.
       8.1.1  If analysis is to be delayed for more than a few hours, adjust the sample pH to less than 2
             with HC1 or H2SO4 solution (Section 7.2) at the time of collection, and refrigerate at 0-4°C
             (40 CFR 136, Table II). To establish the volume of HC1 or H2SO4 required, collect a
             separate aliquot, adjust the pH of this aliquot to less than 2 with acid, and add the volume of
             acid determined to each sample bottle prior to collection. Do not dip pH paper, a pH
             electrode, a stirring rod, or other materials into a sample that will be used for HEM or SGT-
             HEM determination because substances in the sample may adhere to these materials.
       8.1.2  If a sample is known or suspected to contain greater than 1000 mg/L of extractable material,
             collect a proportionately smaller volume of sample (the volume required will depend upon
             the estimated amount of extractable material) in a glass bottle.  Add a proportionately
             smaller amount of HC1 or H2SO4 solution to the smaller sample if preservation is necessary.
 8.2   Collect an additional two aliquots (1 L, additional smaller volume, or both) of a sample for each set
      often samples or less for the matrix spike and matrix spike duplicate.

   Note:    For those circumstances requiring the collection of multiple aliquots of one sample, each
   aliquot is to be collected in either of the following ways: 1) collect  the total volume needed in one
   container, homogenize by mixing with a stir bar until a vortex forms, and siphon into individual
   bottles using a glass tube, 2) collect simultaneously in parallel, if possible, or 3) collect as grab
   samples in rapid succession.

-------
                                                                                       Method 1664
 8.3   The high probability that extractable matter may adhere to sampling equipment and result in
      measurements that are biased low precludes the collection of composite samples for determination
      of oil and grease. Therefore, samples must be collected as grab samples.  If a composite measure-
      ment is required, individual grab samples collected at prescribed time intervals must be analyzed
      separately and the concentrations averaged.
 8.4   All samples must be refrigerated at 0~4°C from the time of collection until extraction (40 CFR 136,
      Table II).
 8.5   All samples must be analyzed within 28 days of the date and time of collection (40 CFR 136, Table
      II).

9.0   Quality Control

 9.1   Each laboratory that uses this method is required to operate a formal quality assurance program
      (Reference 16.8). The minimum requirements of this program consist of an initial demonstration of
      laboratory capability, ongoing analyses of standards and blanks as a test of continued performance,
      and analyses of matrix spike (MS) and matrix spike duplicate (MSD) samples to assess accuracy
      and precision.  Laboratory performance is compared to established performance criteria to deter-
      mine if the results of analyses meet the performance characteristics  of the method.
       9.1.1  The analyst shall make an initial demonstration of the ability to generate acceptable accura-
             cy and precision with this method. This ability is established as described in Section 9.2.
       9.1.2  In recognition of advances that are occurring in analytical technology, the analyst is permit-
             ted certain options to improve separations or lower the  costs of measurements, provided that
             all performance specifications are met. These options include alternate extraction and
             concentration devices and procedures such as solid-phase extraction and Kuderna-Danish
             concentration. Alternate determinative techniques, such as infra-red spectroscopy or
             immuno-assay, and changes that degrade method performance, are not allowed. If an
             analytical technique other than the techniques specified in this method is used, that tech-
             nique must have a specificity equal to or better than the specificity of the techniques in this
             method for HEM and/or SGT-HEM in the sample of interest.  Specificity is  defined as
             producing results equivalent to the results produced by  this method for analytical standards
             (Section  9.2.2) and, where applicable, environmental samples (Section 9.2.3), and that meet
             all of the QC criteria stated in this method.
             9.1.2.1  Each time a modification is made to this method, the analyst is required to repeat
                    the IPR test in Section 9.2.2 to demonstrate that the  modification produces results
                    equivalent to or better than results produced by this  method. If the detection limit
                    of the method will be affected by the modification, the analyst must  demonstrate
                    that the MDL (40 CFR 136, Appendix B) is less than or equal to the MDL in this
                    method or one-third the regulatory compliance  level, whichever is higher. If the
                    modified method is to be used for compliance monitoring, the discharger must also
                    demonstrate that the modified method recovers an amount of HEM and/or SGT-
                    HEM equivalent to the amount recovered by this method on each specific dis-

-------
Method 1664
                    charge. The tests required for this equivalency demonstration are given in Section
                    9.2.3.
            9.1.2.2  The laboratory is required to maintain records of modifications made to this
                    method.  These records include the following, at a minimum:
                    9.1.2.2.1  The names, titles, addresses, and telephone numbers of the analyst(s) who
                             performed the analyses and modification, and of the quality control offi-
                             cer who witnessed and will verify the analyses and modification.
                    9.1.2.2.2  A listing of pollutant(s) measured (HEM and/or SGT-HEM).
                    9.1.2.2.3  A narrative stating reason(s) for the modification.
                    9.1.2.2.4  Results from all quality control (QC) tests comparing the modified
                             method to this method, including:
                             (a)  Calibration (Section 10).
                             (b)  Calibration verification (Section 9.5).
                             (c)  Initial precision and recovery (Section 9.2.2).
                             (d)  Analysis of blanks (Section 9.4).
                             (e)  Accuracy assessment (Section 9.3).
                             (f)  Ongoing precision and recovery (Section 9.6).
                    9.1.2.2.5  Data that will allow an independent reviewer to validate each determina-
                             tion by tracing the instrument output (weight or other signal) to the final
                             result.  These data are to include:
                             (a)  Sample numbers and other identifiers.
                             (b)  Extraction dates.
                             (c)  Analysis dates and times.
                             (d)  Analysis sequence/run chronology.
                             (e)  Sample weight or volume (Section 11.1.4).
                             (f)  Extract volume for SGT-HEM (Section 11.5.2).
                             (g)  Make and model of analytical balance and weights traceable to
                                 NIST.
                             (h)  Copies of logbooks, printer tapes, and other recordings of raw data.
                             (i)  Data system outputs, and other data to link the raw data to the results
                                 reported.
      9.1.3  Analyses of matrix spike and matrix spike duplicate samples are required to demonstrate
            method accuracy and precision and to monitor matrix interferences (interferences caused by
            the sample matrix).  The procedure and QC criteria  for spiking are described in Section 9.3.
      9.1.4  Analyses of laboratory blanks are required to demonstrate freedom from contamination.
            The procedure and criteria for analysis of a blank are described in Section 9.4.
      9.1.5  The laboratory shall, on an ongoing basis, demonstrate  through calibration verification and
            analysis of the ongoing precision and recovery sample that the analysis system is in control.
            These procedures are described in Sections 9.5 and 9.6, respectively.

-------
                                                                                    Method 1664
       9.1.6 The laboratory should maintain records to define the quality of data that is generated.
            Development of accuracy statements is described in Sections 9.3.7 and 9.6.3.
       9.1.7 Accompanying QC for the determination of HEM and/or SGT-HEM is required per analyti-
            cal batch. An analytical batch is a set of samples extracted at the same time, to a maximum
            of 10 samples.  Each analytical batch of 10  or fewer samples must be accompanied by a
            laboratory blank (Section 9.4), an ongoing precision and recovery sample (OPR, Section
            9.6), and a matrix spike and matrix spike duplicate (MS/MSD, Section 9.3), resulting in a
            minimum of five analyses (1 sample, 1 blank, 1  OPR, 1 MS, and 1 MSD) and a maximum
            of 14 analyses (10 samples, 1 blank, 1 OPR, 1 MS, and 1 MSD) in the batch. If greater than
            10 samples are to be extracted at one time, the samples must be separated into analytical
            batches of 10 or fewer samples.
9.2   Initial demonstration of laboratory capability.
       9.2.1 Method Detection Limit (MDL)—To establish the ability to detect HEM and SGT-HEM,
            the analyst shall determine the MDL per the procedure in 40 CFR 136, Appendix B using
            the apparatus, reagents, and standards that will be used in the practice of this method.  An
            MDL less than or equal to the MDL in Section 1.6 must be achieved prior to the practice of
            this method.
       9.2.2 Initial precision and recovery (IPR)—To establish the ability to generate acceptable preci-
            sion and accuracy, the analyst shall perform the  following operations:
            9.2.2.1   Extract and evaporate four samples of the PAR standard (Section  7.11)  according to
                    the procedure beginning in Section  11.
            9.2.2.2  Using the results of the set of four analyses, compute the average  percent recovery
                    (X) and the standard deviation  of the percent recovery  (s) for HEM and for SGT-
                    HEM (if determined). Use the following equation for calculation of the standard
                    deviation of the percent recovery:


                                          Equation 1
                                                   ff*)2
                                               n  - I
                               where:
                               n  = Number of samples
                               x  - Concentration in each sample
            9.2.2.3  Compare s and X with the corresponding limits for initial precision and recovery in
                    Table 1.  If s and X meet the acceptance criteria, system performance is acceptable
                    and analysis of samples may begin. If, however, s exceeds the precision limit or X

-------
 Method 1664
                    falls outside the range for recovery, system performance is unacceptable. In this
                    event correct the problem, and repeat the test.
       9.2.3  Equivalency demonstration for application of a method modification to compliance moni-
             toring—To establish the ability of a modification of this method to recover an amount of
             HEM and/or SGT-HEM equivalent to the amount recovered by this method from a specific
             discharge, proceed as follows:
             9.2.3.1  Collect, extract, concentrate, and weigh the HEM or SGT-HEM in two sets of four
                    aliquots of unspiked wastewater. One set of four wastewater aliquots is analyzed
                    according to the protocol in Section 11 of this method and the other set of four
                    aliquots is analyzed using the modified method.
             9.2.3.2  Calculate the average percent recovery of HEM and SGT-HEM for the set of results
                    from this method and for the set of results from the modified method.  The average
                    percent recovery using the modified method must be 79 to 114 percent of the aver-
                    age percent recovery produced by this method for HEM and  66 to 114 percent of
                    the average percent recovery produced by this method for SGT-HEM. If not, the
                    modified method may not be used.

   Note:    If the average concentration of the four results produced using this method and the
   average concentration of the four results produced using the modified method are below the
   Minimum Level (Section 1.6), and if the equivalency test of the modified method is passed for spikes
   of reference standards into reagent water (Section 9.2.2), the modified method is deemed to be
   equivalent to this method for determining HEM and or SGT-HEM on that specific discharge.

9.3  Matrix spikes—The laboratory must spike, in duplicate, a minimum of 10 percent of all samples
     (one sample in each batch often samples) from  a given sampling site or, if for compliance monitor-
     ing, from a given discharge. The two sample aliquots shall be spiked with the hexadecane/stearic
     acid spiking solution (Section 7.10).
       9.3.1  The concentration of the spike in the sample shall be determined as follows:
            9.3.1.1  If, as in compliance monitoring, the concentration of HEM or SGT-HEM in the
                    sample is being checked against a regulatory concentration limit, the spiking level
                    shall be at that limit or at I to 5 times higher than the background concentration of
                    the sample (determined in Section 9.3.2), whichever concentration is higher.
            9.3.1.2  If the concentration of HEM or SGT-HEM in a sample is not being checked against
                    a limit, the spike shall be at the concentration of the precision and recovery standard
                    (Section 7.11) or at 1 to 5 times  higher than the background concentration, which-
                    ever concentration is higher.
       9.3.2 Analyze one sample aliquot out of each  set of ten samples from each  site or discharge
            according to the procedure beginning in Section 11 to determine the background concen-
            tration (B) of HEM or SGT-HEM
10

-------
                                                                                Method 1664
      9.3.2.1  If necessary, prepare a standard solution appropriate to produce a level in the sam-
              ple at the regulatory compliance limit or at 1 to 5 times the background concentra-
              tion (per Section 9.3.1).
      9.3.2.2  Spike two additional sample aliquots with the spiking solution and analyze these
              aliquots to determine the concentration after spiking (A).
9.3.3  Calculate the percent recovery (P) of HEM or SGT-HEM in each aliquot using the follow-
      ing equation:


                                    Equation 2

                                  p =  100 (A -  B)
                                            7
           where:
           A  = Measured concentration  of analyte after spiking
           B  = Measured background concentration of HEM or SGT-HEM
           T  = True  concentration of the spike (40 mgIL)
      When determining SGT-HEM, the true concentration (T) must be divided by 2 to reflect the
      concentration of hexadecane that remains after removal of stearic acid (20 mg/L).
9.3.4  Compare the percent recovery of the HEM or SGT-HEM with the corresponding QC
      acceptance criteria in Table 1.
      9.3.4.1 If the results of the spike fail the acceptance criteria, and the recovery of the QC
             standard in the ongoing precision and recovery test (Section 9.6) for the analytical
             batch is within the acceptance criteria in Table 1, an interference is present.  In this
             case, the result may not be reported for regulatory compliance purposes and the
             analyst must assess the potential cause for the interference. If the interference is
             attributable to sampling, the site or discharge should be resampled.  If the interfer-
             ence is attributable to a method deficiency, the analyst must modify the method,
             repeat the tests required in Section 9.1.2, and repeat the analysis of the sample and
             the MS/MSD.
      9.3.4.2 If the results of both  the spike and the ongoing precision  and recovery test fail the
             acceptance criteria, the analytical system is judged to be out of control, and the
             problem shall be identified and  corrected, and the sample reanalyzed.
9.3.5  Compute the relative percent difference (RPD) between the two results (not between the
      two recoveries) using the following equation:
                                                                                         11

-------
 Method 1664
                                           Equation 3

                                              D-D.
                                   RPD =	!	L_  x 100
                                           (D, + D2)/2

            D, = Concentration of HEM or SGT-HEM  in the  sample
            D2 = Concentration of HEM or SGT-HEM  in the  second (duplicate) sample
       9.3.6  The relative percent difference for duplicates shall meet the acceptance criteria in Table 1.
             If the criteria are not met, the analytical system is judged to be out of control, and the
             problem must be immediately identified and corrected, and the analytical batch reanalyzed.
       9.3.7  As part of the QC program for the laboratory, method precision and accuracy for samples
             should be assessed and records should be maintained. After the analysis of five spiked
             samples in which the recovery passes the test in Section 9.3.4, compute the average percent
             recovery (Pa) and the standard deviation of the percent recovery (sp). Express the accuracy
             assessment as a percent recovery interval from Pa - 2sp to Pa + 2sp. For example, if Pa =
             90% and sp = 10% for  five analyses of HEM or SGT-HEM, the accuracy interval is ex-
             pressed as 70-110%'. Update the accuracy assessment on a regular basis (e.g., after each
             five to ten new accuracy measurements).
9.4  Laboratory blanks—Laboratory reagent  water blanks are analyzed  to demonstrate freedom from
     contamination.
       9.4.1  Extract and concentrate a laboratory reagent water blank initially (i.e. with the tests in
             Section 9.2) and with each analytical batch.  The blank must be subjected to the same
             procedural steps as a sample.
       9.4.2  If material is detected in the blank at a concentration greater than the Minimum Level
             (Section 1.6), analysis  of samples is halted until the source  of contamination is eliminated
             and a blank shows no evidence of contamination. All samples must be associated with an
             uncontaminated method blank before the results may be reported for regulatory compliance
             purposes.
9.5  Calibration verification—Verify calibration of the balance per Section 10 before and after each
     analytical batch of 14 or fewer measurements.  (The 14 measurements will normally be 10 samples,
     1 blank, 1 OPR, 1 MS, and 1 MSD.) If calibration is not verified after the measurements, recali-
     brate the balance and re weigh the batch.
9.6  Ongoing precision and recovery—To demonstrate that the analysis system is in control, and
     acceptable precision and accuracy is being maintained with each analytical batch, the analyst shall
     perform the following operations:
       9.6.1  Extract and concentrate a precision and recovery standard (Section 7.11) with each  analyti-
             cal batch according to the procedure beginning in Section 11.
12

-------
                                                                                       Method 1664
         9.6.2  Compare the concentration with the limits for ongoing precision and recovery in Table 1.  If
               the concentration is in the range specified, the extraction, evaporation, and weighing pro-
               cesses are in control and analysis of blanks and samples may proceed. If, however, the
               concentration is not in the specified range, the analytical process is not in control.  In this
               event, correct the problem, re-extract the analytical batch, and repeat the ongoing precision
               and recovery test.
         9.6.3  The laboratory should add results that pass the specification in Section 9.6.2 to IPR and
               previous OPR data and update QC charts to form a graphic representation of continued
               laboratory performance. The laboratory should also develop a statement of laboratory data
               quality for each analyte by calculating the average percent recovery (R) and the standard
               deviation of the percent recovery (sr). Express the accuracy as a recovery interval from
               R - 2sr to R + 2sr.  For example, if R = 95% and sr = 5%, the accuracy is 85% to 105%.
  9.7  Quality control sample (QCS)—It is suggested that the laboratory obtain a quality control sample
       from a source different from the source for the hexadecane and stearic acid used routinely in this
       method (Sections 7.8 and 7.9), and that the QCS be used for verification of the concentrations of
       HEM and SGT-HEM using the procedure given in the note in Section 7.10.3.
  9.8  The specifications contained in this method can be met if the apparatus used is scrupulously cleaned
       and dedicated for the determination of HEM and SGT-HEM.  The standards used for initial preci-
       sion and recovery (IPR, Section 9.2.2), matrix spikes (MS/MSD, Section 9.3), and ongoing preci-
       sion and recovery (OPR, Section 9.6) should be identical, so that the most precise results will be
       obtained.
  9.9  Depending upon specific program requirements, field replicates and field spikes of the analytes of
       interest into samples may be required to assess the precision and accuracy of the sampling and
       sample transporting techniques.

10.0  Calibration and Standardization

 10.1   Calibrate the analytical balance at 2 mg and 1000  mg using class "S" weights.
 10.2  Calibration shall be within ± 10%  (i.e. ±0.2 mg) at 2 mg and ± 0.5% (i.e. ±5 mg) at 1000 mg. If
       values are not within these limits, recalibrate the balance.

11.0  Procedure
  This method is entirely empirical.  Precise and accurate  results can be obtained only by strict adherence to
  all details.

     Note:    The procedure below is based on the preparation, extraction, and analysis of a 1 L sample
     volume. If a smaller volume of sample is collected for analysis, the laboratory may need to adjust
     the size of the labware  used in order to compensate for the smaller volume being processed.
     Commensurately smaller volumes of reagents (i.e. HCl or H2SO4, n-hexane, and sodium sulfate)
     may be used.
                                                                                                13

-------
  Method 1664
 11.1   Preparation of the analytical batch.
       11.1.1  Bring the analytical batch of samples, including the sample aliquots for the MS and MSD,
              to room temperature.
       11.1.2  Place approximately 1000 mL (950-1050 mL) of reagent water (Section 7.1) in a clean
              sample bottle to serve as the laboratory blank.
       11.1.3  Prepare the OPR (Section 9.6) using the PAR standard (Section 7.11).
       11.1.4  Either mark the sample bottle at the water meniscus or weigh the bottle for later deter-
              mination of sample volume. Weighing will be more accurate. Mark or weigh the MS and
              MSD.
 11.2   pH verification.
       11.2.1  Verify that the pH of the sample is less than 2 using the following procedure:
             11.2.1.1  Dip a glass stirring rod into the well mixed sample.
             11.2.1.2  Withdraw the stirring rod and allow a drop of the sample to fall on or touch the pH
                     paper.

    Note:   Do not dip the pH paper into the bottle or touch it to the sample on the lid.

             11.2.1.3  Rinse the stirring rod with a small portion  of n-hexane that will be used for extrac-
                     tion (to ensure that no extractable material is lost on the stirring rod). Collect the
                     rinsate in the separatory funnel to be used for sample extraction.
       11.2.2  If the sample is at neutral pH, add 5-6 mL of HC1 or H2SO4 solution (Section 7.2) to the 1
              liter sample. If the sample  is at high pH, use a proportionately larger amount of HC1 or
              H2SO4 solution. If a smaller sample volume was collected, use a proportionately smaller
              amount of HC1 or H2SO4 solution
       11.2.3  Replace the cap and shake the bottle to mix thoroughly. Check the pH of the sample using
              the procedure in Section 11.2.1. If necessary,  add  more acid to the sample and retest.
       11.2.4  Add the appropriate amount of HC1 or H2SO4 solution to the blank, OPR, MS, and MSD.
11.3  Extraction
       11.3.1  Tare a clean boiling flask containing 3-5 boiling chips as follows:
            11.3.1.1   Place the  flask containing the chips in  an oven at  105-115°C for a minimum of 2 h
                     to dry the flask and chips.
            11.3.1.2  Remove from the oven and immediately transfer to cool to room temperature in a
                     desiccator.
            11.3.1.3  When cool, remove from the desiccator with tongs and weigh immediately on a
                     calibrated balance (Section 10).
       11.3.2 Pour the sample into the separatory funnel.
 14

-------
                                                                                       Method 1664
       11.3.3  Add 30 mL n-hexane to the sample bottle and seal the bottle with the original bottle cap.
              Shake the bottle to rinse all interior surfaces of the bottle, including the lid of the bottle cap.
              Pour the solvent into the separatory funnel.
       1 1 .3.4  Extract the sample by shaking the separatory funnel vigorously for 2 minutes with periodic
              venting into a hood to release excess pressure.
       11.3.5  Allow the organic phase to separate from the aqueous phase for a minimum of 10 minutes.
              If an emulsion forms between the phases and the emulsion is greater than one-third the
              volume of the solvent layer, the analyst must employ mechanical techniques to complete the
              phase separation.  The optimum technique depends upon the sample, but may include
              stirring, filtration through glass wool, use of solvent phase separation paper, centrifugation,
              use of an ultrasonic bath with ice, addition of NaCl, or other physical methods.  Alternative-
              ly, solid-phase, continuous, or other extraction techniques may be used to prevent emulsion
              formation, provided that the requirements in Section 9.1.2 are  met.
       1 1 .3.6  Drain the aqueous layer (lower layer) into the original sample  container. Drain a small
              amount of the organic layer into the sample container to minimize the amount of water
              remaining in the separatory funnel.

    Note:    The amount of water remaining with the n-hexane must be minimized to prevent dissolu-
    tion or clumping of the sodium  sulfate in the solution drying process.
       11.3.7  Place approximately 10 g anhydrous NajSC^ in a filter funnel and rinse with a small portion
              of n-hexane. Discard the rinsate.

    Note:    The specific properties of a sample may necessitate the use of larger amounts ofNa2SO4.
       1 1 .3.8  Drain the n-hexane layer (upper layer) from the separatory funnel through the NajSC^ into
              the preweighed boiling flask containing the boiling chips (Section 11.3.1.3).
       1 1 .3.9  Repeat the extraction (Sections 1 1 .3.3-1 1 .3.6 and 1 1 .3.8) twice more with fresh 30-mL
              portions of n-hexane, combining the extracts in the boiling flask.
      11.3.10  Rinse the tip of the separatory funnel, the filter paper, and the funnel with 2-3 small (3-5
              mL) portions of n-hexane.  Collect the rinsings in the flask.
      11.3.11  A milky extract  indicates the presence of water. If the extract is milky, allow the solution to
              stand for up to one hour to allow the water to settle. Decant the solvent layer (upper layer)
              through sodium  sulfate to remove any excess water as  in Sections 1 1.3.7  and 1 1 .3.8.
      1 1 .3.1 2  If only SGT-HEM is to be determined, proceed to Section 11.5.
1 1 .4  Solvent evaporation.
       11.4.1  Connect the boiling flask to the distilling head apparatus and evaporate the solvent by
              immersing the lower half of the flask in a water bath or a steam bath. Adjust the water
              temperature as required to complete the concentration  in less than 30 minutes.  Collect the
              solvent for reuse.

                                                                                                15

-------
  Method 1664
       11.4.2  When the temperature in the distilling head reaches 70°C or the flask appears almost dry,
              remove the distilling head. Sweep out the flask for 15 seconds with air to remove solvent
              vapor by inserting a glass tube connected to a vacuum source.  Using tongs, immediately
              remove the flask from the heat source and wipe the outside surface dry to remove moisture
              and fingerprints.

    Note:   The analyst should carefully monitor the flask during the final stages of evaporation to
    assure that all of the solvent is removed and at the same time to prevent loss of the more volatile
    sample constituents.

       11.4.3  Inspect the residue in the boiling flask for crystals. Crystal formation is an indication that
              sodium sulfate may have dissolved and passed into the boiling flask. This may happen if
              the drying capacity of the sodium sulfate is exceeded or if the sample is not adjusted to a
              low pH. If crystals are observed, redissolve the extract in n-hexane, filter into another tared
              boiling flask, and repeat the evaporation procedure (Sections 11.4.1-11.4.2).
       11.4.4  Cool the boiling flask to room temperature in a desiccator and maintain in the desiccator for
              30 minutes minimum.  Remove with tongs and weigh immediately to determine the weight
              of the material in the flask.
            11.4.4.1  If the extract was from the HEM procedure, determine the HEM (Wh) by subtract-
                     ing the tare weight (Section  11.3.1) from the total weight of the flask.
            11.4.4.2  If the extract was from the SGT-HEM procedure (Section 11.5.5), determine the
                     weight of SGT-HEM (Ws) by subtracting the tare weight from the total weight of
                     the flask.
       11.4.5  Determine the original sample volume (Vs) in liters by filling the sample bottle to the mark
              with water and measuring the volume of water in a 1 to 2 L graduated cylinder.  If the
              sample weight was used (Section 11.1.4), weigh the empty bottle and cap and  determine Vs
              by difference,  assuming a sample density of 1.00.
11.5   SGT-HEM determination.
       11.5.1   Silica gel capacity—To ensure that the capacity of the silica gel will not be exceeded, the
              amount of HEM must be less than 100 mg or, if above 100 mg, must be known.
            11.5.1.1  If it  is  known that the amount of HEM is  less than 100 mg, the analyst may proceed
                     with the determination of SGT-HEM per Sections 11.5.3-11.5.5 without determina-
                     tion  of HEM.
            11.5.1.2  If, however, the amount of HEM is not known, HEM must first be determined using
                     the procedure in Sections 11.3-11.4.
      11.5.2   Extractable materials in silica gel—Because the capacity of silica gel is not known for all
              substances, it is presumed that 3 g will adsorb 100 mg of all adsorbable materials. The
              amount of silica gel that can be used for adsorption in the SGT-HEM procedure below has
              been limited to 30 g because of concerns about possible extractable impurities  in the silica
 16

-------
                                                                                       Method 1664
               gel. Therefore, if the extract contains more than 1000 mg of HEM, split the extract per the
               following procedure:
             11.5.2.1  Add 85-90 mL of n-hexane to the boiling flask to redissolve the HEM. If neces-
                      sary, heat the solution on an explosion-proof hotplate or in a water bath to com-
                      pletely redissolve the HEM.
             11.5.2.2  Transfer the extract to a 100-mL volumetric flask.  Rinse the boiling flask sequen-
                      tially with 2-3 small portions of n-hexane and add to the volumetric flask.  Dilute to
                      the mark with n-hexane.
             11.5.2.3  Calculate the extract volume that contains 1000 mg of extractable material accord-
                      ing to  the following equation:


                                            Equation 4

                                                 1000 V
                                           V  =	1
                    where:
                     Va - Volume of aliquot to be withdrawn (mL)
                     Vt = Total volume of solvent used in Section 11.5.2.2 (mL)
                    Wh - Weight of extractable material  HEM  measurement (mg)
             11.5.2.4  Using a calibrated pipet, remove the volume to be withdrawn (Va) and return to the
                      boiling flask.  Dilute to approximately 100 mL with n-hexane.
       11.5.3  Adsorption with silica gel
             11.5.3.1  Add 3.0 ± 0.3 g of anhydrous silica gel (Section 7.7) to the boiling flask for every
                      100 mg of HEM, or fraction thereof, to a maximum of 30 g of silica gel.  For exam-
                      ple, if the weight of HEM is 735 mg, add 3 x 8 = 24 g of silica gel.
             11.5.3.2  Add a fluoropolymer-coated stirring bar to the flask and stir the solution on a mag-
                      netic stirrer for a minimum of 5 minutes.
       11.5.4  Filter the solution through n-hexane moistened filter paper into a pre-dried, tared boiling
              flask containing several boiling chips. Rinse the silica gel and filter paper with several
              small amounts of n-hexane to complete the transfer.
       11.5.5  Evaporate the solution and determine the weight of SGT-HEM per Section 11.4.

12.0   Data Analysis and Calculations

 12.1   n-Hexane extractable material—Calculate the concentration of HEM ("oil and grease") in the
       sample per the following equation:
                                                                                                17

-------
   Method 1664
                                            Equation 5

                                                      W. (mg)
                                      HEM (mg/L) =    h   *

                      where:
                      Wh = Weight of extractable material Section 11.4.4.1 (mg)
                      V = Sample volume from Section 11.4.5 (L)
  12.2   Silica gel treated n-hexane extractable material—Calculate the concentration of SGT-HEM ("petro-
        leum hydrocarbons1') in the sample per the equation above, substituting Ws (from Section 11.4.4.2)
        for Wh. If the extract was split to decrease the total amount of material to 1,000 mg, determine the
        corrected total weight of SGT-HEM in the un-split extract (Wc) using the following equation:


                                            Equation 6


                                       Wc  (mg) = _i Wd (mg)
                                                   a
     where:
     Wd = Weight in the portion of the extract split for adsorption (Sections 11.5.2.4 and 11.4.4.2)
     Vt and Va are as defined in Equation 4
       Use the corrected total weight of SGT-HEM in the unsplit extract (Wc) to determine the total SGT-
       HEM in the sample by substituting Wc for Wh in Equation 5.
 12.3  Reporting

       12.3.1   Samples—Report results to three significant figures for HEM and SGT-HEM found above
               the ML (Section 1.6) in all samples. Report results below the ML as < 5 mg/L for HEM
               and SGT-HEM, or as required by the permitting authority or permit.
       12.3.2   Report results to three significant figures for HEM and SGT-HEM found above the MDL
               (Section  1.6) in all blanks.  Do not report results below the MDL unless required by the
               permitting authority or in the permit.

13.0   Method Performance

       This method was validated through single laboratory studies and an inter-laboratory method valida-
       tion study (Reference 16.9).
  18

-------
                                                                                     Method 1664
14.0   Pollution Prevention

 14.1   The solvents used in this method pose little threat to the environment when recycled and managed
       properly.
 14.2   Standards should be prepared in volumes consistent with laboratory use to minimize the volume of
       expired standards to be disposed.

15.0   Waste Management

 15.1   It is the laboratory's responsibility to comply with all federal, state, and local regulations governing
       waste management, particularly the hazardous waste identification rules and land disposal restric-
       tions, and to protect the air, water, and land by minimizing and controlling all releases from fume
       hoods and bench operations. Compliance with all sewage discharge permits and regulations is also
       required.
 15.2   Samples preserved with HC1 or H2SO4 to pH < 2 are hazardous and must be neutralized before
       being disposed, or must be handled as hazardous waste.
 15.3   For further information on waste management, consult "The Waste Management Manual for
       Laboratory Personnel", and "Less is Better: Laboratory Chemical Management for Waste Reduc-
       tion", both available from the American Chemical Society's Department of Government Relations
       and Science Policy, 1155 16th Street N.W., Washington, D.C. 20036.

16.0   References

 16.1   "Methods for Chemical Analysis of Water and Wastes", 3rd Edition, Environmental Protection
       Agency, Environmental Monitoring Systems Laboratory-Cincinnati (EMSL-Ci), Cincinnati, Ohio
       45268, EPA-600/4-79-020, Method 413.1, (1983).
 16.2   "Standard Methods for the Examination of Water and Wastewater", 18th Edition, American Public
       Health Association, 1015 Fifteenth Street, NW, Washington, D.C. 20005, Method 5520B and
       Method 5520F, (1992).
 16.3   40 CFR 136, Appendix A, Methods 1624 and 1625.
 16.4   "Carcinogens - Working With Carcinogens," Department of Health, Education, and Welfare, Public
       Health Service, Center for Disease Control, National Institute for Occupational Safety and Health,
       Publication No. 77-206, August 1977.
 16.5   "OSHA Safety and Health Standards, General Industry," (29 CFR 1910), Occupational Safety and
       Health Administration, OSHA 2206 (Revised, January 1976).
 16.6   "Safety in Academic Chemistry Laboratories," American Chemical Society, Committee on Chemi-
       cal Safety, 3rd  Edition, 1979.
 16.7   "Standard Practices for Sampling Water," ASTM Annual Book of Standards, Part 31, D3370-76,
       American Society for Testing and Materials, 1916 Race Street, Philadelphia, PA 19103-1187, 1980.
 16.8   "Handbook of Analytical Quality Control in Water and Wastewater Laboratories," USEPA, EMSL-
       Ci, Cincinnati,  OH 45268, EPA-600/4-79-019, March 1979.
                                                                                             19

-------
 Method 1664
16.9   Report of the Method 1664 Validation Studies, April 1995. Available from the Sample Control
      Center (operated by DynCorp Environmental Programs Division), 300 N. Lee Street, Alexandria,
      VA 22314, (703)519-1140.
20

-------
                                                                                             Method 1664
17.0   Tables
  Table 1. Acceptance Criteria for Performance Tests
     Acceptance Criterion
Section
Limit (%)
     Initial precision and recovery
         HEM Precision (s)
         HEM Recovery (X)
         SGT-HEM Precision (s)
         SGT-HEM Recovery (X)
    9.2.2
  9.2.2.2
  9.2.2.2
  9.2.2.2
  9.2.2.2
          10
     83-101
          13
     83-116
     Matrix spike/matrix spike duplicate
         HEM Recovery
         HEM RPD
         SGT-HEM Recovery
         SGT-HEM RPD
     9.3
    9.3.4
    9.3.5
    9.3.4
    9.3.5
      79-114
          18
      66-114
          24
     Ongoing precision and recovery
         HEM Recovery
         SGT-HEM Recovery
     9.6
     9.6
     9.6
     79-114
     66-114
                                                                                                       21

-------
  Method 1664
18.0   Glossary of Definitions and Purposes
       The definitions and purposes are specific to this method but have been conformed to common usage
       as much as possible.
 18.1   Units of weight and measure and their abbreviations
       18.1.1  Symbols
              °C            degrees Celsius
              <             less than
              %            percent
              ±             plus or minus
       18.1.2  Alphabetical characters
              g             gram
              h             hour
              L             liter
              mg           milligram
              mg/g          milligram per gram
              mg/L          milligram per liter
              mg/mL        milligram per milliliter
              mL           milliliter
              No.           number
              rpm          revolutions per minute
 18.2   Definitions, acronyms, and abbreviations
       18.2.1  Analyte: The HEM or SGT-HEM tested for by this method.
       18.2.2  Analytical batch: The set of samples extracted at the same time, to a maximum of 10
              samples. Each analytical batch of 10 or fewer samples must be accompanied by a laborato-
              ry blank (Section 9.4), an ongoing precision and recovery sample (OPR, Section 9.6), and a
              matrix spike and matrix spike duplicate (MS/MSD, Section 9.3), resulting in a minimum of
              five analyses (1 sample, 1 blank, 1 OPR, 1 MS, and 1 MSD) and a maximum of 14 analyses
              (10 samples,  1 blank, 1 OPR, 1 MS, and 1  MSD) in the batch. If greater than 10 samples
              are to be extracted  at one time, the samples must be separated into analytical batches of 10
              or fewer samples.
       18.2.3  Field blank: An aliquot of reagent water that is placed in a sample container in the labora-
              tory or in the field and treated as a sample  in all respects, including exposure to sampling
              site conditions, storage, preservation, and all analytical procedures. The purpose of the
              field blank is to determine if the field or sample transporting procedures and environments
              have contaminated the sample.
       18.2.4  HEM:  See n-Hexane extractable material.
       18-2.5  n-Hexane extractable material: The material that is extracted from a sample and determined
              by this method.
       18.2.6  IPR: See initial precision and  recovery.

 22

-------
                                                                                 Method 1664
 18.2.7  Initial precision and recovery (IPR): Four aliquots of the diluted PAR analyzed to establish
        the ability to generate acceptable precision and accuracy. An IPR is performed the first
        time this method is used and any time the method or instrumentation is modified.
 18.2.8  Laboratory blank (method blank): An aliquot of reagent water that is treated exactly as a
        sample including exposure to all glassware, equipment, solvents, reagents, internal stan-
        dards, and surrogates that are used with samples.  The laboratory blank is used to determine
        if analytes or interferences are present in the laboratory environment, the reagents, or the
        apparatus.
 18.2.9  Laboratory control sample (LCS): See Ongoing precision and recovery standard (OPR).
18.2.10  Matrix spike (MS) and matrix spike duplicate (MSP): Aliquots of an environmental
        sample to which known quantities of the analytes are added in the laboratory. The MS and
        MSD are prepared and/or analyzed exactly like a field sample.  Their purpose is to quantify
        any additional bias and imprecision caused by the sample matrix.  The background concen-
        trations of the  analytes in the sample matrix must be determined in a separate aliquot and
        the measured values in the MS and MSD corrected for background concentrations.
18.2.11  May: This action,  activity, or procedural step is neither required nor prohibited.
18.2.12  May not: This action, activity, or procedural step is prohibited.
18.2.13  Method Detection Limit:  The lowest level at which an analyte can be detected with 99
        percent confidence that the analyte concentration is greater than zero.
18.2.14  Minimum Level (ML): The lowest level at which the entire analytical system gives a
        recognizable signal and acceptable calibration point for the analyte.  It is equivalent to the
        concentration of the lowest calibration standard, assuming that all method-specified sample
        weights, volumes, and cleanup procedures have been employed.
18.2.15  Must: This action, activity, or procedural step is required.
18.2.16  Ongoing precision  and recovery standard (OPR, also called a laboratory control sample): A
        laboratory blank spiked with known quantities of analytes. The OPR is analyzed exactly
        like a sample.  Its purpose is to assure that the results produced by the laboratory remain
        within the limits specified in this method for precision and accuracy.
18.2.17  OPR: See Ongoing precision and recovery standard.
18.2.18  PAR: See Precision and recovery standard.
18.2.19  Precision and recovery standard: Secondary standard that is diluted and spiked to form the
        IPR and OPR.
18.2.20  Quality control sample (QCS): A sample containing analytes of interest at known concen-
        trations. The QCS is obtained from a source external to the laboratory or is prepared from
        standards obtained from a different source than the calibration standards. The purpose is to
        check laboratory performance using test materials that have been prepared independently
        from the normal preparation process.
18.2.21  Reagent water: Water demonstrated to be free from HEM and SGT-HEM and potentially
        interfering substances at or above the Minimum Level of this method.
                                                                                          23

-------
 Method 1664
      18.2.22  SGT-HEM:  See Silica gel treated n-hexane extractable material.
      18.2.23  Should:  This action, activity, or procedural step is suggested but not required.
      18.2.24  Silica gel treated n-hexane extractable material:  n-Hexane extractable material (HEM) that
              is not adsorbed by silica gel.
      18.2.25  Stock solution:  A solution containing an analyte that is prepared using a reference material
              traceable to EPA, the National Institute of Science and Technology (NIST), or a source that
              will attest to the purity and authenticity of the reference material.
24

-------

-------

-------