©EPA  Method 1632: Determination of
       Inorganic Arsenic in Water by
       Hydride Generation Flame Atomic
       Absorption
                                   ) Printed on Recycled Paper

-------
Method 1632
                                   Acknowledgments
Method 1632 was prepared under the direction of William A. Telliard of the  U.S. Environmental
Protection Agency's (EPA's) Office of Water (OW), Engineering and Analysis Division (BAD).  The
method was prepared under EPA Contract 68-C3-0337 by the DynCorp Environmental Programs Division
with assistance from Quality Works,  Inc. and Interface, Inc.  The method is based on procedures
developed by Eric Crecelius of the Battelle Marine Sciences Laboratory hi Sequim, Washington.
                                        Disclaimer
This method has been reviewed and approved for publication by the Engineering and Analysis Division
of the U.S. Environmental Protection Agency. Mention of trade names or commercial products does not:
constitute endorsement or recommendation for use.
 Questions concerning this method or its application should be addressed to:

 W.A. Telliard
 USEPA Office of Water
 Analytical Methods Staff
 Mail Code 4303
 401 M Street, SW
 Washington, DC 20460
 Phone: 202/260-7120
 Fax:  202/260-7185
 ii
                                                                           Draft, January 1996

-------
                                                                                     Method 1632
                                         Introduction

 This analytical method was designed to support water quality monitoring programs authorized under the
 Clean Water Act.  Section 304(a) of the Clean Water Act requires EPA to publish water quality criteria
 that reflect the latest scientific knowledge about the physical fate (e.g., concentration and dispersal) of
 pollutants, the effects of pollutants on ecological and human health, and the effect of pollutants on
 biological community diversity, productivity, and stability.

 Section 303 of the Clean Water Act requires states to set a water quality standard for each body of water
 within its boundaries. A state water quality standard consists of a designated use or uses of a waterbody
 or a segment of a waterbody, the water quality criteria that are necessary to protect the designated use or
 uses, and an antidegradation policy. These water quality standards serve two purposes: (1) they establish
 the water quality goals for a specific waterbody, and (2) they are the basis for establishing water quality-
 based treatment controls and strategies beyond the technology-based controls required by Sections 301(b)
 and 306 of the Clean Water Act.

 In defining water quality standards, the state may use narrative criteria, numeric criteria, or both.
 However, the 1987 amendments to the Clean Water Act required states to adopt numeric criteria for toxic
 pollutants (designated in Section  307(a) of the Act) based  on EPA Section  304(a) criteria or other
 scientific data, when the discharge or presence of those toxic pollutants could reasonably  be expected to
 interfere with designated uses.

 In some cases, these water quality criteria are as much as 280 times lower than those achievable using
 existing EPA methods and required to support technology-based permits. Therefore, EPA developed new
 sampling and analysis methods to specifically address state needs for measuring toxic metals at water
 quality criteria levels, when such measurements are necessary to protect designated uses in  state water
 quality standards. The latest criteria published by EPA are those listed in the National Toxics Rule (57
 PR 60848) and the Stay of Federal Water Quality Criteria for Metals (60 FR 22228).  These rules include
 water quality criteria for 13  metals, and it is these criteria on which the new sampling and analysis
 methods are  based.   Method 1632 was specifically developed to provide reliable  measurements of
 inorganic arsenic at EPA WQC levels using hydride generation flame atomic absorption techniques.

 In developing these methods, EPA found that one of the greatest difficulties in measuring pollutants at
 these levels was precluding sample contamination during collection, transport, and analysis. The degree
 of difficulty, however, is highly dependent on the metal and site-specific conditions. This analytical
method, therefore, is designed to provide the level of protection necessary to preclude contamination in
nearly all situations.   It is also designed to provide the procedures necessary to produce reliable results
 at the lowest possible water quality criteria published by EPA. In recognition of the variety of situations
to which this method may be applied, and in recognition of continuing technological advances, the method
is •performance based. Alternative procedures may be used, so long as those procedures are demonstrated
to yield reliable results.

Requests for additional copies should be directed to:

U.S. EPA NCEPI
 11029 Kenwood Road
Cincinnati, OH 45242
513/489-8190

-------
Method 1632
    Note:  This method is intended to be performance based, and the laboratory is permitted to omit
    any step or modify any procedure if all performance requirements set forth hi this method are
    met The laboratory is not allowed to omit any quality control analyses. The terms "must,"
    "may," and "should" are included throughout this method aiad are intended to illustrate the
    importance of the procedures hi producing verifiable data at water quality criteria levels. The
    term "must" is used to indicate that researchers hi trace metals analysis have found  certain
    procedures essential hi successfully analyzing samples and avoiding contamination;  however,
    these procedures can be modified or omitted if the laboratory can demonstrate that data quality
    is not affected.
 iv
                                                                                Draft, January 1996

-------
                                                                              Method 1632
                               Method  1632
                      Inorganic Arsenic in Water by
                      Hydride Generation Flame AA
1.0   Scope and Application

1.1    This method is for determination of total inorganic arsenic (As) in filtered and unfiltered water
       by hydride generation and flame atomic absorption detection. The method is for use in EPA's
       data gathering and monitoring programs associated with the Clean Water Act. The method is
       based on a contractor-developed method (Reference 16.1) and on peer-reviewed, published
       procedures for the determination of arsenic in aqueous samples (Reference 16.2).

1.2    This method is accompanied by Method 1669:  Sampling Ambient Water for Determination of
       Trace Metals at EPA Water Quality Criteria Levels (Sampling Method).  The Sampling
       Method is necessary to ensure that contamination will not compromise trace metals
       determinations during the sampling process.

1.3    This method is designed for measurement of dissolved and total arsenic in the range of 10-200
       ng/L. This method is not intended for determination of arsenic at concentrations normally
       found in treated and untreated discharges from industrial facilities.  Existing regulations (40
       CFR Parts 400-500) typically limit concentrations in industrial discharges to the part-per-
       billion (ppb) range, whereas ambient arsenic concentrations are normally hi the low part-per-
       trillion (ppt) range.

1.4    The detection limits and quantitation levels in this method are usually dependent on the level
       of background elements rather than instrumental limitations. The method detection limit
       (MDL; 40 CFR 136, Appendix B) for total inorganic arsenic has been determined to be 2 ng/L
       when no background elements or interferences are present. The minimum level (ML) has been
       established at 10 ng/L.

1.5    The ease of contaminating water samples with the metal(s) of interest and interfering
       substances cannot be overemphasized. This method includes  suggestions for improvements hi
       facilities and analytical techniques that should maximize the ability of the laboratory to make
       reliable trace metals determinations and minimize contamination. Section 4.0  gives these
       suggestions. Additional suggestions for improvement of existing facilities may be found in
       EPA's Guidance for Establishing Trace Metals Clean Rooms in Existing Facilities, which is
       available.from the National Center for Environmental Publications and Information (NCEPI) at
       the address listed in the introduction to this document.
Draft, January 1996

-------
Method 1632
1.6    Clean and ultraclean—The terms "clean" and "ultraclean" have been applied to the techniques
       needed to reduce or eliminate contamination in trace metals determinations. These terms are
       not used in this method because they lack an exact definition.  However, the information
       provided in this method is consistent with EPA's summiuy guidance on clean and ultraclean
       techniques.

1.7    This method follows the EPA Environmental Methods Management Council's "Format for
       Method Documentation."

1.8    This method is "performance based." The analyst is permitted to modify the method to
       overcome interferences or lower the cost of measurements if all performance criteria are met.
       Section 9.1.2 gives the requirements for establishing method equivalency.

1.9    Any modification of this method, beyond those expressly permitted, shall be considered a
       major modification subject to application and approval of alternate test procedures at 40 CFR
       136.4 and 136.5.

1.10   Each analyst who uses this method must demonstrate the ability to  generate acceptable results
       using the procedure in Section 9.2.

1.11   This method is accompanied by a data verification and validation guidance document,
       Guidance on the Documentation and Evaluation of Trace Metals Data Collected for CWA
       Compliance Monitoring.  Before this method is used, data users should state data quality
       objectives (DQOs) required for a project.


2.0   Summary of Method

2.1    A 100-2000 mL sample is collected directly into a specially cleaned, pretested, fluoropolymer,
       conventional or linear polyethylene, polycarbonate, or polypropylene bottle using sample
       handling techniques specially designed for collection of metals at trace levels (Reference 16.3).

2.2    The sample is either field or laboratory preserved by the addition of 5 mL of pretested 10%
       HNO3 per liter of sample, depending on the time between sample collection and arrival at the
       laboratory.
 2.3    An aliquot of sample is removed and placed in a specially designed reaction vessel.

 2.4
Before analysis, 6 M HC1 and 4% NaBH4 solution are added to convert organic and inorganic
arsenic to volatile arsines.
 2.5    The arsines are purged from the sample onto a cooled glass trap packed with 15% OV-3 on
        Chromasorb® WAW-DMCSO, or equivalent.
                                                                             Draft, January 1996

-------
                                                                                    Method 1632
 2.6    The trapped arsines are thermally desorbed, in order of increasing boiling points, into an inert
        gas stream that carries them into the flame of an atomic absorption spectrophotometer for
        detection.  The first arsine to be desorbed will be AsH3, which represents total inorganic
        arsenic in the sample.

 2.7    Quality is ensured through calibration and testing of the hydride generation, purging, and
        detection systems.
 3.0   Definitions

 3.1     As defined by this method, total inorganic arsenic means all NaBH4-reducible inorganic arsenic
        compounds found in aqueous solution. This method can be extended to measure organic
        arsenic as well. Organic arsenic includes, but is not limited to, monomethylarsonate and
        dimethylarsonate.  In this context, "total" arsenic refers to the forms and species of arsenic, not
        to the total recoverable or dissolved fraction normally determined in an unfiltered or filtered
        sample, respectively. In this method, the total recoverable fraction will be referred to as "total
        recoverable" or "unfiltered."

 3.2     As defined by this method, dissolved inorganic arsenic means all NaBH4-reducible inorganic
        arsenic compounds found in aqueous  solution filtrate after passing the sample through a 0.45-
        uM capsule filter.

 3.3     Definitions of other  terms used in this method are given in the glossary at the end of the
        method.
4.0    Contamination and Interferences

4.1     Preventing ambient water samples from becoming contaminated during the sampling and
        analytical process constitutes one of the greatest difficulties encountered in trace metals
        determinations.  Over the last two decades, marine chemists have come to recognize that much
        of the historical data on the concentrations of dissolved trace metals in seawater are
        erroneously high because the concentrations reflect contamination from sampling and analysis
        rather than ambient levels. Therefore, it is imperative that extreme care be taken to avoid
        contamination when collecting and analyzing ambient water samples for trace metals.

4.2     Samples may become contaminated by numerous routes.  Potential sources of trace metals
        contamination during sampling include:  metallic or metal-containing labware (e.g., talc gloves
        that contain high levels of zinc), containers, sampling equipment, reagents, and reagent water;
        improperly cleaned and stored equipment, labware, and reagents; and atmospheric inputs such
        as dirt and dust.  Even human contact can be a source of trace metals contamination.
Draft, January 1996

-------
Method 1632
4.3    Contamination Control

       4.3.1   Philosophy—The philosophy behind contamination control is to ensure that any object
               or substance that contacts the sample is metal fr
-------
                                                                                    Method 1632
        4.3.6   Wear gloves—Sampling personnel must wear clean, nontalc gloves during all
               operations involving handling of the apparatus, samples, and blanks.  Only clean
               gloves may touch the apparatus.  If another object or substance is touched, the glove(s)
               must be changed before again handling the apparatus.  If it is even suspected that
               gloves have become contaminated, work must be halted, the contaminated gloves
               removed, and a new pair of clean gloves put on.  Wearing multiple layers of clean
               gloves will allow the old pair to be quickly stripped with minimal  disruption to the
               work activity.

        4.3.7   Use metal-free apparatus—All apparatus used for determination of metals at ambient
               water quality criteria levels must be nonmetallic, free of material that may contain
               metals, or both.

               4.3.7.1 Construction materials—Only fluoropolymer (FEP, FIFE), conventional or
                      linear polyethylene, polycarbonate, or polypropylene containers should be used
                      for samples that will be analyzed for arsenic.  PTFE is less desirable than FEP
                      because the sintered material in PTFE may contain contaminants and is
                      susceptible to serious memory effects (Reference 16.4). All materials,
                      regardless of construction, that will directly or indirectly contact the sample
                      must be cleaned using the procedures given in this method and must be known
                      to be clean and metal free before proceeding.

               4.3.7.2 Serialization—It is recommended that serial numbers be indelibly marked or
                      etched on each piece of apparatus so that contamination can be traced,  and
                      logbooks should be maintained to track the sample from the container through
                      the labware to injection into the instrument. It may be useful to dedicate
                      separate sets of labware to different sample types; e.g., receiving waters vs.
                      effluents.  However, the apparatus used for processing blanks and standards
                      must be mixed with the apparatus used to process samples  so that
                      contamination of all labware can be detected.

               4.3.7.3 The laboratory or cleaning facility is responsible for cleaning the apparatus
                      used by the sampling team. If there are any indications that the apparatus is
                      not clean when received by the sampling team (e.g., ripped storage bags), an
                      assessment of the likelihood of contamination must be made. Sampling must
                      not proceed if it is possible that the apparatus is contaminated.  If the
                      apparatus is contaminated, it must be returned to the laboratory or cleaning
                      facility for proper cleaning before any sampling activity resumes.

        4.3.8   Avoid sources of contamination—Avoid contamination by being aware of potential
               sources and routes of contamination.

               4.3.8.1 Contamination by carryover—Contamination may occur when a sample
                      containing low concentrations of metals is processed immediately after a
Draft, January 1996

-------
Method 1632
                      sample containing relatively high concentrations of these metals. To reduce
                      carryover, the sample introduction system may be rinsed between samples with
                      dilute acid and reagent water.  When an unusually concentrated sample is
                      encountered, it is followed by analysis of a laboratory blank to check for
                      carryover.  Samples known or suspected to contain the lowest concentration of
                      metals should be analyzed first followed by samples containing higher levels.

               4.3.8.2 Contamination by samples—Significant laboratory or instrument contamination
                      may result when untreated effluents, in-pirocess waters, landfill leachates, and
                      other samples containing high concentrations of inorganic substances are
                      processed and analyzed.  This method is not intended for application to these
                      samples, and samples containing high concentrations should not be permitted
                      into the clean room and laboratory dedicated for processing trace metals
                      samples.

               4.3.8.3 Contamination by indirect contact—apparatus that may not directly come in
                      contact with the samples may still be a source of  contarnination. For example,
                      clean tubing placed hi a dirty plastic bag may pick up contamination from the
                      bag and subsequently transfer the contamination to the sample.  Therefore, it is
                      imperative that every piece of the apparatus that is directly or indirectly used
                      in the collection, processing, and analysis of ambient water samples be cleaned
                      as specified in Section 11.

               4.3.8.4 Contamination by airborne paniculate matter—Less obvious substances capable
                      of contaminating samples include airborne particles.  Samples may be
                      contaminated by airborne dust, dirt, particles, or vapors  from unfiltered air
                      supplies; nearby corroded or rusted pipes, wires, or other fixtures; or metal-
                      containing paint.  Whenever possible, sample processing and analysis should
                      occur as far as possible from sources of airborne  contamination.
 4.4 Interferences


 4.4.1   If the transfer line between the cold trap and the atomizer is not well heated, water vapor may
        condense in the line.  Such condensation can interfere with the determination of
        dimethylarsine.

 4.4.2   High concentrations of cobalt, copper, iron, mercury, or nickel can cause interferences through
        precipitation as reduced metals and blockage of transfer lines and fittings
                                                                               Draft, January 1996

-------
                                                                                      Method 1632
 5.0   Safety

 5.1     The toxicity or carcinogenicity of each chemical used in this method has not been precisely
        determined; however, each compound should be treated as a potential health hazard.  Exposure
        to these compounds should be reduced to the lowest possible level.  It is recommended that
        the laboratory purchase a dilute standard solution of the As in this method. If primary
        solutions are prepared, they shall be prepared in a hood, and a NIOSH/MESA-approved toxic
        gas respirator shall be worn when high concentrations are handled.

 5.2     This method does not address all safety issues associated with its use. The laboratory is
        responsible for maintaining a current awareness file of OSHA regulations for the safe handling
        of the chemicals specified in this method. A reference file of material safety data sheets
        (MSDSs) should also be made available to all personnel involved in these analyses. It is also
        suggested that the laboratory perform personal hygiene monitoring of each analyst who uses
        this method and that the results of this monitoring be made available to the analyst. Additional
        information on laboratory safety can be found in References 16.5-16.8. The references and
        bibliography at the end of Reference 16.8 are particularly comprehensive in dealing with the
        general subject of laboratory safety.

 5.3     Samples suspected to contain high concentrations of As are handled using essentially the same
        techniques used in handling radioactive or infectious materials. Well-ventilated, controlled
        access laboratories are required. Assistance in evaluating the health hazards of particular
        laboratory conditions may be  obtained from certain consulting laboratories and from State
        Departments of Health or Labor, many of which have an industrial health service. Each
        laboratory must develop a strict safety program for handling As.

        5.3.1   Facility—When samples known or suspected of containing high concentrations of
               arsenic are handled, all operations (including removal of samples from sample
               containers, weighing, transferring, and mixing) should be performed in a glove box
               demonstrated to be leaktight or in a fume hood demonstrated to have adequate air
               flow. Gross losses to the laboratory ventilation system must not be allowed. Handling
               of the dilute solutions normally used in analytical and animal work presents no
               inhalation hazards except in an accident.

               Protective equipment—Disposable plastic gloves, apron or laboratory coat, safety
               glasses or mask, and a glove box or fume hood adequate for radioactive work should
               be used. During analytical operations that may give rise to aerosols or dusts, personnel
               should wear respirators equipped with activated carbon filters.

        5.3.3   Training—Workers must be trained in the proper method of removing contaminated
               gloves and clothing without contacting the exterior surfaces.
5.3.2
Draft, January 1996

-------
Method 1632
       5.3.4   Personal hygiene—Hands and forearms should te washed thoroughly after each
               manipulation and before breaks (coffee, lunch, and shift).

       5.3.5   Confinement—Isolated work areas posted with signs, segregated glassware and tools,
               and plastic absorbent paper on bench tops will aid in confining contamination.

       5.3.6   Effluent vapors—The effluent from the AAS should pass through either a column of
               activated charcoal or a trap designed to remove As.

       5.3.7   Waste handling—Good technique includes minimizJng contaminated waste.  Plastic
               bag liners should be used in waste cans. Janitors and other personnel must be trained
               in the safe handling of waste.

       5.3.8   Decontamination

               5.3.8.1 Decontamination of personnel—Use any mild soap with plenty of scrubbing
                      action.

               5.3.8.2 Glassware, tools, and surfaces—Satisfactory cleaning may be accomplished by
                      washing with any detergent and water.

       5.3.9   Laundry—Clothing known to be contaminated should be collected in plastic bags.
               Persons who convey the bags and launder the clothing should be advised of the hazard
               and trained hi proper handling. If the launderer knows of the potential problem, the
               clothing may be put into a washer without contact. The washer should be run through
               a cycle before being used again for other clothing.

        5.3.10  Wipe tests—A useful method of determining clisanliness of work surfaces and tools is
               to wipe the surface with a piece of filter paper. Extraction and analysis by this method
               can achieve a limit of detection of less than 1 ng per wipe. Less than 0.1 ug per wipe
               indicates acceptable cleanliness; anything higher warrants further cleaning. More than
               10 ug on a wipe constitutes an acute hazard and requires prompt cleaning before
               further use of the equipment or work space, and indicates that unacceptable work
               practices have been employed.
                                                          I

 6.0   Apparatus and Materials

        Note: Brand names, suppliers, and part numbers are for illustration purposes only
        and no endorsement is  implied.  Equivalent performance may be achieved using
        apparatus and materials other than those specified here.  The laboratory is responsible
        for meeting the performance requirements of this  method.      	    '
 8
                                                                              Draft, January 1996

-------
                                                                                     Method 1632
 6.1     Sampling equipment

         6.1.1   Sample collection botties-Fluoropolymer, conventional or linear polyethylene,
                polycarbonate, or polypropylene, 500-1000 mL

         6.1.2   Cleaning—Bottles are cleaned with liquid detergent and thoroughly rinsed with reagent
                water. They are then heated to 50-60°C in concentrated reagent grade HNO3 for at
                least 2 h. The bottles are rinsed with reagent water, then immersed in a hot (50-60°)
                bath of IN trace metal grade HC1 for at least 48 h.  The bottles are then thoroughly
                rinsed with reagent water and filled with 0.1% (v/v) ultrapure HC1 and double-bagged
                in new polyethylene zip-type bags until needed.

 6.2    Equipment for bottle and glassware cleaning

        6.2.1   Vats, up to 200-L capacity, constructed of high-density polyethylene (HDPE) or other
                nonmetallic, noncontaminating material suitable for holding concentrated HNO3 and
                dilute HC1

        6.2.2   Panel immersion heater, all-fluoropolymer coated, capable of maintaining a
                temperature of 60-75°C in a vat of the size used

        6.2.3   Laboratory sink in class 100 clean area, with high-flow reagent water for rinsing

        6.2.4   Clean bench, class 100, for drying rinsed bottles.

        Atomic absorption spectrophotometer (AAS)—Any atomic absorption spectrophotometer may
        serve as a detector. A bracket is required to hold the quartz atomizer in the optical path of the
        instrument.  Table 1 gives typical conditions for the spectrophotometer.

        6.3.1   Electrodeless  discharge lamp for measuring arsenic at 193.7 nm

        6.3.2  Quartz cuvette burner tube (Reference 16.2), 70 mm long and 9 mm in diameter with
               two 6-mm o.d. side tubes, each 25 mm long. Figure la shows a schematic of the  tube
               and bracket

        Equipment for reaction vessel—Figure Ib shows the schematic diagram for the vessel used for
        the reaction of the sample with sodium borohydride.  The system consists of the following:

        6.4.1   125-mL gas wash bottle (Corning #  1760-125 or equivalent) onto which an 8-mm  o.d.
               sidearm inlet tube 2 cm long has been grafted

        6.4.2   Silicone rubber stopper septum (Ace Glass #9096-32 or equivalent)
6.3
6.4
Draft, January 1996

-------
Method 1632
        6.4.3   Four-way Teflon stopcock valve capable of switching of the helium from the purge to
               the analysis mode of operation

        6.4.4   Flow meter/needle valve capable of controlling ;md measuring gas flow rate to the
               reaction vessel at 150 (± 30) mL/min

        6.4.5   Silicone tubing—All glass-to-glass connections ;are made with silicone rubber sleeves.

6.5     Equipment for cryogenic trap—Figure Ic shows the schematic diagram for the  trap. It consists
        of the following:

        6.5.1   Nichrome wire (22-gauge)

        6.5.2   Variacs for controlling Nichrome wire

        6.5.3   A 6-mm o.d. borosilicate glass U-tube about 30 cm long with a 2-cm radius of bend
            .   (or similar dimensions to fit into a tall widemouth Dewar flask), which has been
               silanized and half-packed with 15% OV-3 on Cboromasorb® WAW-DMCS (45-60
               mesh), and the ends of which have been packed with silanized glass wool.

               6.5.3.1  Conditioning the trap—The input side of the trap (the side that is not packed)
                       is connected with silicone rubber tubing; to He at a flow rate of 40 mL/min,
                       and the trap is placed in an oven at 175°C for 2 h. At the end of this time,
                       inject two 25-uL aliquots of GC column conditioner (Silyl-8®, Supelco, Inc., or
                       equivalent) through the silicone tubing into the glass trap. Return the trap to
                       the oven, with the He still flowing, for 24 h.

               6.5.3.2 After conditioning, the trap is wrapped with approximately  1.8 m of 22-gauge
                       Nichrome wire, the ends of which are affixed to crimp-on electrical contacts.
                       The wire-wrapped column is coated with approximately 2 mm of silicon
                       rubber caulking compound.

               6.5.3.3 The trap is connected by silicone rubber tubing to the output of the reaction
                       vessel. The output side of the trap is connected by 6-mm diameter borosilicate
                       tubing that has been wrapped by Nichrome wire to the input of the flame
                       atomizer.

        6.5.4  Dewar flask capable of containing the trap described in Section 6.5.3.

 6.6    Recorder/integrator—Any integrator with a range compatible with the AAS is  acceptable.

 6.7    Pipettors—All-plastic pneumatic fixed volume  and variable pipettors hi the range of 10 uL to
        5.0 mL.
 10
                                                                               Draft, January 1996

-------
                                                                                   Method 1632
6.8    Analytical balance capable of weighing to the nearest 0.01 gram


7.0   Reagents and Standards

7.1    Reagent water—Water demonstrated to be free from As and potentially interfering substances
       at the MDL. Prepared by distillation, deionization, reverse osmosis, anodic/cathodic stripping
       voltammetry, or other technique that removes As and potential interferant(s).

7.2    Hydrochloric acid—Trace-metal purified reagent HC1.

7.3    6M hydrochloric acid—Equal volumes of trace metal grade concentrated HC1 (Section 7.2) and
       reagent water (Section 7.1) are combined to give a solution approximately 6M in HC1.

7.4    Nitric acid—Trace-metal purified reagent HNO3

7.5    10% nitric acid—Nitric acid (HNO3):  10% wt, Seastar, or equivalent.

7.6    Sodium borohydride solution-4 g of >98% NaBH4 (previously analyzed and shown to be free
       of measurable arsenic) are dissolved in  100 mL of 0.02 M NaOH solution. This solution is
       stable for only 8-10 h, and must be made fresh daily.

7.7    Liquid nitrogen for cooling the cryogenic trap

7.8    Helium.—Grade  4.5 (standard laboratory grade) helium.

7.9    Hydrogen—Grade 4.5 (standard laboratory grade) hydrogen.

7.10   Air—Grade 4.5 (standard laboratory grade) air.

7.11   QC sample concentrate—From a source different than the one used to prepare calibration
       standards, prepare a QC sample concentrate containing 1.50 ug/L As in reagent water (Section
       7.1) preserved with sufficient HNO3 to bring the pH to <2.

7.12   Arsenic Standards

       7.12.1   Stock standard solution (1000 mg/L)—Either procure A2LA-certified aqueous
             •  standards from a supplier and verify by comparison to a second standard from a
               separate source, or dissolve 1.320 g of arsenic trioxide (As^) in 100 mL of reagent
               water (Section 7.1) containing 4 g NaOH.  Acidify the solution with 20 mL
               concentrated HNO3 and dilute to 1.0 liter.
Draft, January 1996
11

-------
Method 1632
       7.12.2 Primary dilution standard (1.0 mg/L)—Pipet 1.0 mL arsenic stock solution (Section
              7.12.1) into a 1000-mL volumetric flask and bring to volume with reagent water
              (Section 7.1) containing 1.5 mL concentrated HNO3 per liter.

       7.12.3 Standard arsenic solution (10 ug/L)—Pipet 10.0 itnL primary dilution arsenic standard
              (Section 7.12.2) into a 1000-mL volumetric flask, and bring to volume with reagent
              water (Section 7.1) containing 1.5 mL concentrated HNO3/liter (1 mL = 0.01 ug As).

       7.12.4 Calibration solutions—Using the standard arsenic:  solution (Section 7.12.3), prepare
              calibration solutions, one of which contains As at a concentration of 10 ng/L and at
              least two others that fall within the calibration range of the instrument in reagent water
              (Section 7.1).


8.0   Sample Collection, Preservation, and Storage

8.1    Sample collection—Samples are collected as described in the Sampling Method (Reference
       16.3).

8.2    Sample filtration—For dissolved metals,  samples and field blanks are filtered through a 0.45-
       uM capsule filter at the field site. The Sampling Method describes the filtering procedures.

8.3    Sample preservation—Preservation of samples may be performed in the field or in the
       laboratory. The sampling team may prefer to use laboratory preservation of samples to
       expedite field operations and to mirdmize the potential for sample contamination.  Samples and
       field blanks should be preserved at the laboratory immediately when received.  Preservation
       involves the addition of 10% HNO3 (Section 7.1.3) to bring the sample to pH < 2.  For
       samples received at neutral pH, approximately 5 mL of 10% HNO3 per liter will be required.

       8.3.1   Wearing clean gloves, remove the cap from the sample bottle, add the volume of
              reagent grade acid that will bring the pH to <2, :and recap the bottle immediately. If
              the bottle is full, withdraw the necessary volume; using a precleaned pipet and then add
              the acid. Record the volume withdrawn and the amount of acid used.
        Note: Do not dip pH paper or a pH meter into the saraiple; remove a small aliquot
        •with a clean pipet and test the aliquot.	
        8.3.2   Store the preserved sample for a minimum of 48 h at 0-4°C to allow the acid to
               completely dissolve the metal(s) adsorbed on the container walls.
 12
Draft, January 1996

-------
                                                                                     Method 1632
        8.3.3   With each sample set, preserve a method blank and an OPR sample in the same way
                as the sample(s).

        8.3.4   Sample bottles should be stored in polyethylene bags at 0-4°C until analysis.


 9.0   Quality Control/Quality Assurance

 9.1     Each laboratory that uses this method is required to operate a formal quality assurance
        program (Reference 16.3).  The minimum requirements of this program consist of an initial
        demonstration of laboratory capability, analysis of samples spiked with As to evaluate and
        document data quality, and analysis of standards and blanks as  tests of continued performance.
        To determine if the results of analyses meet the performance characteristics of the method,
        laboratory performance is compared to established performance criteria.

        9.1.1    The analyst shall make an initial demonstration of the ability to generate acceptable
                accuracy and precision with this method.  This ability is established as described in
                Section 9.2.

        9.1.2    In recognition of advances that are occurring in analytical technology, the analyst is
                permitted to exercise certain options to eliminate interferences or lower the costs of
                measurements. These options include alternate digestion, concentration, and cleanup
                procedures, and changes in instrumentation. Alternate determinative techniques such
                as the substitution of a colorimetric technique or changes  that degrade method
                performance are not allowed. If an analytical technique other than the techniques
                specified in this method is used, that technique must have a specificity equal to or
                better than the specificity of the techniques in the referenced method for the analytes
                of interest.

                9.1.2.1 Each time this method is modified, the analyst is required to repeat the
                      procedure in Section 9.2.  If the change will affect the detection limit of the
                      method, the laboratory is required to demonstrate  that the MDL (40 CFR Part
                      136, Appendix B) is lower than the MDL for this method or one-third of the
                      regulatory compliance level, whichever is higher.  If the change will affect
                      calibration, the analyst must recalibrate the instrument according to Section 10
                      of this method.

               9.1.2.2 The laboratory is required to maintain records of modifications made to this
                      method.  These records include the following, at a minimum:

                      9.1.2.2.1       The names, titles,  addresses, and telephone numbers of the
                                     analyst(s) who performed the analyses and modification, and of
                                     the quality control officer who witnessed and will verify the
                                     analyses and modification
Draft, January 1996
13

-------
Method 1632
                      9.1.2.2.2       A listing of metal measured (As), by name and CAS Registry
                                     number

                      9.1.2.2.3       A narrative stating reasons) for the modification(s)

                      9.1.2.2.4       Results from all quality control (QC) tests comparing the
                                     modified method to this method, including:

                                     (a)  Calibration
                                     (b)  Calibration verification
                                     (c)  Initial precision and recovery (Section 9.2)
                                     (d)  Analysis of blanks
                                     (e)  Accuracy assessment

                      9.1.2.2.5       Data that will allow an independent reviewer to validate each
                                     determination by tracing the instrument output (peak height,
                                     area, or other signal) to ithe final result.  These data are to
                                     include, where possible:

                                     (a)  Sample numbers and other identifiers
                                     (b)  Preparation dates
                                     (c)  Analysis dates and times
                                     (d)  Analysis sequence/nun chronology
                                     (e)  Sample volume
                                     (f)  Volume before each preparation step
                                     (g)  Volume after each preparation step
                                     (h)  Final volume before analysis
                                     (i)  Dilution data
                                     (j)  Instrument and operating conditions (make, model,
                                             revision, modifications)
                                     (k)  Sample introduction system (ultrasonic nebulizer, hydride
                                             generator, flow injection system, etc.)
                                     (1)  Operating conditions (ashing temperature, temperature
                                             program, flow rates, etc.)
                                     (m) Detector (type, operating conditions, etc.)
                                     (n)  Printer tapes and other recordings of raw data
                                     (o)  Quantitation reports, data system outputs, and other data to
                                             link the raw data to the results reported

        9.1.3   Analyses of blanks are required to demonstrate freedom from contamination.  Section
               9.5 describes the required types, procedures, and criteria for analysis of blanks.

        9.1.4   The laboratory shall spike at least 10% of the ssimples with As to monitor method
               performance.  Section 9.3 describes this test.  ^Tien results of these spikes indicate
 14
Draft, January 1996

-------
                                                                                     Method 1632
               atypical method performance for samples, an alternative extraction or cleanup
               technique must be used to bring method performance within acceptable limits.  If
               method performance for spikes cannot be brought within the limits given in this
               method, the result may not be reported for regulatory compliance purposes.

        9.1.5   The laboratory shall, on an ongoing basis, demonstrate through calibration verification
               and through analysis of the ongoing precision and recovery aliquot that the analytical
               system is  in control. Sections 9.6 and 10.2 describe these procedures.

        9.1.6   The laboratory shall maintain records to define the quality of data that are generated.
               Section 9.3.4 describes the development of accuracy statements.

 9.2     Initial demonstration of laboratory capability

        9.2.1   Method detection limit—To establish the ability to detect As, the analyst shall
               determine the MDL per the procedure in 40 CFR 136, Appendix B using the
               apparatus, reagents, and standards that will be used in the practice of this method.  The
               laboratory must produce an MDL that is no more than one-tenth the  regulatory
               compliance level or that is less than the MDL listed in Table 1, whichever is greater.

        9.2.2   Initial precision and recovery (IPR)—To establish the ability to generate acceptable
               precision and recovery, the analyst shall perform the following operations.

               9.2.2.1 Analyze four aliquots of reagent water spiked with As at 0.015 ug/L according
                      to the procedures in this method.  Prepare these by diluting each of four 1.0-
                      mL aliquots of the QC Sample Concentrate to 100 mL.  All digestion,
                      extraction, and concentration steps, and the containers, labware, and reagents
                      that will be used with samples must be used in this test.

               9.2.2.2 Using results of the set of four analyses, compute the average percent recovery
                      (X) in each aliquot and the standard deviation(s) of the recovery.

               9.2.2.3 Compare s and X with the corresponding limits  for initial precision and
                      recovery in Table 2.  If s and X meet the acceptance criteria, system
                      performance is acceptable and analysis of blanks and samples may begin.  If,
                      however, s exceeds the precision limit or X falls outside the range for
                      accuracy, system performance is unacceptable for As.  Correct the problem and
                      repeat the test (Section 9.2.2.1).

9.3     Method accuracy—To assess the performance of the method on a given sample matrix, the
        laboratory must perform matrix spike (MS) and matrix spike duplicate (MSD) sample analyses
        on 10% of the samples from each site being  monitored, or at least one MS sample analysis
        and one MSD sample analysis must be performed for each sample set (samples collected from
        the same site at the same time, to a maximum of 10 samples), whichever is more frequent
Draft, January 1996
15

-------
Method 1632
       9.3.1   The concentration of the MS and MSD is determined as follows:

               9.3.1.1 If, as hi compliance monitoring, the concentration of As hi the sample is being
                      checked against a regulatory concentration limit, the spike must be at that limit
                      or at five times the background concentration, whichever is greater.

               9.3.1.2 If the concentration is not being checked against a regulatory limit, the
                      concentration must be at five times the background concentration or at five
                      times the ML in Table 1, whichever is greater.

       9.3.2   Assessing spike recovery

               9.3.2.1 Determine the background concentration (B) of As by analyzing one sample
                      aliquot according to the procedures specified hi this method.

               9.3.2.2 If necessary, prepare a matrix spiking concentrate that will produce the
                      appropriate level (Section 9.3.1) in the ssimple when the concentrate is added.

               9.3.2.3 Spike a second  sample aliquot with the QC check sample concentrate (or,  if
                      necessary, with the matrix spiking concentrate) and analyze it to determine the
                      concentration after spiking (A).

               9.3.2.4 Calculate each percent recovery (P) as 100(A - B)/T, where T is the known
                      true value of the spike.

        9.3.3   Compare the percent recovery (P)  with the corresponding  QC acceptance criteria hi
               Table 2. If P falls outside the designated range for recovery, As has failed the
               acceptance  criteria.

               9.3.3.1  If As has failed the acceptance criteria, analyze the ongoing precision and
                       recovery standard (Section 9.6).  If the OPR is within its respective limit
                       (Table 2), the analytical system is hi control and the problem can be attributed
                       to the sample matrix.

               9.3.3.2 For samples that exhibit matrix problems, further isolate As from the sample
                       matrix using chelation, extraction, concentration, hydride generation, or other
                       means, and repeat the accuracy test (Section 9.3.2).

               9.3.3.3 If the recovery for As remains outside the acceptance criteria, the analytical
                       result in the unspiked sample is suspect and may not be reported for regulatory
                       compliance purposes.

         9.3.4  Recovery for samples should be assessed and records maintained.
 16
                                                                                Draft, January 1996

-------
                                                                                     Method 1632
                9.3.4.1 After the analysis of five samples of a given matrix type (river water, lake
                       water, etc.) for which As passes the tests in Section 9.3.3, compute the average
                       percent recovery (R) and the standard deviation of the percent recovery (SR).
                       Express the accuracy assessment as a percent recovery interval from R - 2SR
                       to R + 2SR for each matrix.  For example, if R = 90% and SR = 10% for five
                       analyses of river water, the accuracy interval is expressed as 70-110%.

                9.3.4.2 Update the accuracy assessment in each matrix regularly (e.g., after each five
                       to ten new measurements).
 9.4    Precision of matrix spike duplicates
        9.4.1
Calculate the relative percent difference (RPD) between the MS and MSD according to
the equation below using the concentrations found in the MS and MSD; Do not use
the recoveries calculated in Section 9.3.2.4 for this calculation because the RPD of
recoveries is inflated when the background concentration is near the spike
concentration.
                                     RPD
                                                (Dl+D2)/2
                      Where:
                      Dl = concentration of the analyte in the MS sample
                      D2 = concentration of the analyte in the MSD sample
        9.4.2   Compare the RPD with the limits in Table 2.  If the criteria are not met, the analytical
               system is judged to be out of control.  Correct the problem and reanalyze all samples
               in the sample set associated with the MS/MSD that failed the RPD test.

9.5     Blanks—Blanks are analyzed to demonstrate freedom from contamination.

        9.5.1   Laboratory (method) blank

               9.5.1.1 Prepare a method blank with each sample batch (samples of the same matrix
                      started through the preparation process on the same 12-hour shift, to a
                      maximum of 10 samples). Analyze the blank immediately after analysis of the
                      OPR (Section 9.6) to demonstrate freedom from contamination.

               9.5.1.2 If As or any potentially interfering  substance is found in the blank at a
                      concentration equal to or greater than the MDL (Table 1), sample analysis
                      must be halted, the source of the contamination determined, the problem
                      corrected, and  the sample batch and fresh method blank reanalyzed.
Draft, January 1996
                                                                                            17

-------
Method 1632
               9.5.1.3 Alternatively, if a sufficient number of blanks (three minimum) are analyzed to
                      characterize the nature of a blank, the average concentration plus two standard
                      deviations must be less than the regulatoiy compliance level.

               9.5.1.4 If the result for a single blank remains atiove the MDL or if the result for the
                      average concentration plus two standard deviations of three or more blanks
                      exceeds the regulatory compliance level, results for samples associated with
                      those blanks may not be reported for regulatory compliance purposes. Stated
                      another way, results for all initial precision and recovery tests (Section 9.2)
                      and all samples must be associated with m uncontaminated method blank
                      before these results may be reported for regulatory compliance purposes.

        9.5.2  Field blank

               9.5.2.1 Analyze the field blank(s) shipped with each set of samples (samples collected
                      from the same site at the same time, to a, maximum of 10 samples).  If the
                       samples are filtered for the determinatioE of dissolved As, the field blank shall
                      be filtered as well.  Analyze the blank immediately before analyzing the
                       samples in the batch.

               9.5.2.2 If As or any potentially interfering substance is found in the field blank at a
                       concentration equal to or greater than the MDL  (Table 1), or greater than one-
                       fifth the level in the associated sample, whichever is greater,  results for
                       associated samples may be the result of contamination and may not be reported
                       for regulatory compliance purposes.

               9.5.2.3 Alternatively, if a sufficient number of field blanks (three minimum) are
                       analyzed to characterize the nature of the field blank, the average concentration
                       plus two standard deviations must be less  than the regulatory compliance level
                       or less than one-half the level in the associated sample, whichever is greater.

               9.5.2.4 If contamination of the field blanks and associated samples is known or
                       suspected, the laboratory should commuioicate this to the sampling team so that
                       the source of contamination can be identified and corrective  measures taken
                       before the next sampling event.

         9.5.3   Equipment blanks—Before any sampling equipment is used at a given site, the
                laboratory or cleaning facility  is required to generate equipment blanks to demonstrate
                that the sampling equipment is free from contamination. Two types of equipment
                blanks are required:  bottle blanks and sampler check blanks.

                9.5.3.1 Bottle blanks—After undergoing appropriate cleaning procedures (Section
                        11.4), bottles should be subjected to conditions of use to verify the
                       effectiveness of the cleaning procedures.  A representative set of sample bottles
  18
                                                                                Draft, January 1996

-------
                                                                                     Method 1632
                       should be filled with reagent water acidified to pH < 2 and allowed to stand
                       for a minimum of 24 h. Ideally, the time that the bottles are allowed to stand
                       should be as close as possible to the actual time that sample will be in contact
                       with the bottle. After standing, the water should be analyzed for any signs of
                       contamination. If any bottle shows signs of contamination, the problem must
                       be identified, the cleaning procedures corrected or cleaning solutions changed,
                       and all affected bottles recleaned.

                9.5.3.2 Sampler check blanks—Sampler check blanks are generated hi the laboratory
                       or at the equipment cleaning contractor's facility by processing reagent water
                       through the sampling devices using the same procedures that are used in the
                       field (see Sampling Method).  Therefore, the "clean hands/dirty hands"
                       technique used during field sampling should be followed when preparing
                       sampler check blanks at the laboratory or cleaning facility.

                       9.5.3.2.1       Sampler check blanks are generated by filling a large carboy or
                                     other container with reagent water (Section 7.2) and processing
                                     the reagent water through the equipment using the same
                                     procedures that are used in the field (see Sampling Method).
                                     For example, manual grab sampler check blanks are collected
                                     by directly submerging a sample bottle into the water, filling
                                     the bottle, and capping.  Subsurface sampler check blanks are
                                     collected by immersing the sampler into the water and
                                     pumping water into a sample container.  "Clean hands/dirty
                                     hands" techniques must be used.

                      9.5.3.2.2       The sampler check blank must be analyzed using the
                                     procedures in this method. If As or any potentially interfering
                                     substance is detected hi the blank, the source of contamination
                                     or interference must be identified and the problem corrected.
                                     The equipment must be demonstrated to be free from  As
                                     before the equipment may be used hi the field.

                      9.5.3.2.3       Sampler check blanks must be run on all equipment that will
                                     be used hi the field.  If, for example, samples are to be
                                     collected using both a grab sampling device and a  subsurface
                                     sampling device, a sampler check blank must be run on both
                                     pieces of equipment.

9.6     Ongoing precision and recovery

        9.6.1   Prepare a precision and recovery sample (laboratory-fortified method blank) identical
               to the initial precision and recovery aliquots  (Section 9.2) with each sample batch
               (samples of the same matrix started through the extraction process on the same 12-
Draft, January 1996
                                                                                             19

-------
Method 1632
              hour shift, to a maximum of 10 samples) by diluting a 1.0-mL aliquot of QC check
              sample concentrate to 100 mL with reagent water.

       9.6.2  Analyze the OPR aliquot before analyzing the method blank and samples from the
              same batch.

       9.6.3  Compute the percent recovery of As hi the OPR aliquot using the procedure given hi
              this method.

       9.6.4  Compare the concentration to the limits for ongoiing recovery hi Table 2.  If the
              acceptance criteria are met, system performance is acceptable and analysis of blanks
              and samples may proceed.  If, however, recovery falls outside of the range given, the
              analytical processes are not being performed properly.  Correct the problem, reprepare
              the sample batch, and repeat the OPR test (Section 9.6).

       9.6.5  Add results that pass the specifications in Section 9.6.4 to initial and previous ongoing
              data for As hi each matrix. Update QC charts to form a graphic representation of
              continued laboratory performance. Develop a statement of laboratory accuracy hi each
              matrix type by calculating the average percent recovery (R) and the standard deviation
              of percent recovery (SR). Express the accuracy as a recovery interval from R - 2SR to
              R + 2SR. For example, if R = 95% and SR = 5%, the accuracy is 85-105%.

 9.7    The specifications hi this method can be met if the instrument used is calibrated properly and
       then maintained hi a calibrated state. A given instrument will provide the most reproducible
       results if dedicated to the settings and conditions required  for the analyses of As by this
       method.                                             |

 9.8    Depending on specific program requirements, field duplicates may be collected to determine
        the precision of the sampling technique. The relative percent difference (RPD) between field
        duplicates should be less than 20%.
 10.0  Calibration  and Standardization

 10.1   Calibration—Calibrate at a minimum of three points, one of which must be the ML (Table 1),
        and another that must be near the upper end of the linear dynamic range. Calibration is
        required before any samples or blanks are analyzed.

        10.1.1   External  standard calibration

                iO.1.1.1        Calculate the response factor (RF) for As hi each CAL solution using
                              the following equation and the height or area produced.
 20
                                                                              Draft, January 1996

-------
                                                                                     Method 1632
                                           RF
                          Where:
                          Rx = height or area of the signal for As
                          Cx = concentration  of compound injected (v-g/L)
                10.1.1.2       Calculate the mean RF (M), the standard deviation of the RF (SD), and
                              the relative standard deviation (RSD) of the mean, where RSD = 100 x
                              SD/M.

                10.1.1.3       Linearity—If the RSD of the mean RF is less than 25% over the
                              calibration range, an averaged response factor may be used.
                              Otherwise, a calibration curve must be used over the calibration range.

        Note:   The reagents used in the hydride generation process are likely to cause some
        absorbance in the absence of As.  Using a mean RF to quantitate samples -will
        therefore result in decreasing accuracy with decreasing concentrations of As. To
        avoid this, it is recommended that a calibration curve that is not forced through the
 	origin be used in lieu of a mean RF, even if the %RSD of the Rfs is less than 25%.

 10.2    Calibration verification—Immediately after calibration, an initial calibration verification should
        be performed.  Adjustment of the instrument is performed until verification criteria are met.
        Only after these criteria are met may blanks and samples be analyzed.

        10.2.1  Verify the specificity of the instrument for As and adjust the wavelength or tuning
               until the resolving power specified in this method is met.

        10.2.2  Inject the mid-point calibration standard (Section 10.1) or a dilution of the QC check
               sample concentrate with a concentration near the midpoint of the calibration range.

        10.2.3  Compute the percent recovery of As using the mean response or calibration curve
               obtained hi the  initial calibration.

        10.2.4  Compare the recovery with the corresponding limit for calibration verification hi Table
               2.  If acceptance criteria are met, system performance is acceptable and analysis  of
               blanks and samples may continue  using the response from the initial calibration.  If
               acceptance criteria are not met, system performance is unacceptable. Locate and
               correct the problem and/or prepare a new calibration check standard and repeat the test
               (Sections 10.2.2-10.2.4), or recalibrate the system according to this method and
               Section  10.1.
Draft, January 1996
                                                                                             21

-------
Method 1632
       10 2 5  Calibration should be verified following every ten samples by analyzing the mid-point
               calibration standard. If the recovery does not meet the acceptance criteria specified in
               Table 2, analysis must be halted, the problem corrected, and the instrument
               recalibrated.  All samples after the last acceptable calibration verification must be
               reanalyzed.


 11.0 Procedure

 11.1   Set the AA system up according to manufacturer's instructions. The settings hi Tables 3 and 4
       can be used as a guide. Calibrate the instrument according to Section 10.

       Note:  Precision and sensitivity are affected by gas flow rates and these must be
       individually optimized for each system, using the figures in Table 4 as an initial guide.

 11 2  To light the flame, turn on all gases and expose the end of the quartz cuvette to a flame.  At
       this point, a flame will be burning out the ends of the tobe.  Allow the tube to heat for
       approximately 5 min, then place a flat metal spatula over each end of the tube in sequence.
       An invisible air/hydrogen flame should now be burning in the center  of the cuvette. To check
       for the flame, place a mirror near the end of the tube.

 113  Place a known volume of aqueous sample (up to 70 mL) into the reaction vessel.  If less than
        70 mL of sample is used, add sufficient reagent water to result in a total volume of 70 mL.
        Add 5.0 mL of 6M HC1.  Set the four-way valve on the reaction vessel to pass the flow of He
        through the sample and onto the trap and begin purging the vessel with He.

 11.4   Lower the trap  into a Dewar flask containing liquid nitrogen (LNj) and top the flask off with
        LN2 to a constant level.

 115   Add 10 mL of NaBH4 solution slowly (over a period of approximately 2 min) through the
        rubber septum with a disposable hypodermic syringe  and begin timing the reaction. After 7
        min, turn the stopcock on the four-way valve to bypass the reaction vessel and pass hehum
        directly to the trap.                                 i

 11.6   Quickly remove the trap from the LN2, activate the heating coils to heat the trap and transfer
        line to 80"C, and begin recording output from the AA system.

 117   To ensure that  all organic reduction products have been desorbed from the trap, maintain the
        trap temperature at 80°C and keep He flowing through the trap for at least 3 min between
        samples.
  22
                                                                              Draft, January 1996

-------
                                                                                   Method 1632
  12.0  Data Analysis and Calculations

  12.1    Compute the concentration of As in ng/L (parts-per-trillion; ppt) using the calibration data
         (Section 10.1) according to the following equation:
                                       C, G«Q -

                          where the terms are defined in Section 10.1.1

 12.2   If the concentration exceeds the linear dynamic range of the instrument, dilute the sample by
        successive factors of 10 until the concentration is within the linear dynamic range.

 12.3   Report results at or above the ML for As found in samples and determined in standards.
        Report all results for As found in blanks, regardless of level.

 12.4   Report results to one significant figure at or below the MDL, two significant figures between
        the MDL and ML, and three significant figures at or above the ML.

 12.5   Do not perform blank subtraction on the sample results.
 13.0  Method  Performance

        Before this method was documented, an MDL study was conducted using the techniques
        described in the method.  Additional method performance studies will be conducted in 1995,
        and the method will be revised as needed to reflect the results of those studies.

 14.0  Pollution Prevention
 14.1
Pollution prevention encompasses any technique that reduces or eliminates the quantity or
toxicity of waste at the point of generation. Many opportunities for pollution prevention exist
in laboratory operation. EPA has established a preferred hierarchy of environmental
management techniques that places pollution prevention as the management option of first
choice.  Whenever feasible, laboratory personnel should use pollution prevention techniques to
address their waste generation. When wastes cannot be feasibly reduced at the source, the
Agency recommends recycling as the next best option.  The acids used in this method should
be reused as practicable by purifying by electrochemical techniques. The only other chemicals
used in this method are the neat materials used in preparing standards. These standards are
used in extremely small amounts and pose little threat to the environment when managed
Draft, January 1996
                                                                                          23

-------
Method 1632
       properly. Standards should be prepared in volumes consistent with laboratory use to minimize
       the disposal of excess volumes of expired standards.

14.2   For irrfbrmation about pollution prevention that may be applied to laboratories and research
       institutions, consult Less is Better: Laboratory Chemical Management for Waste Reduction,
       available from the American Chemical Society's Department of Governmental Relations and
       Science Policy, 1155 16th Street NW, Washington DC 20036, 202/872-4477.
                                                        I

15.0 Waste Management

15.1   The laboratory is responsible for  complying with all federal, slate, and local regulations
       governing waste management, particularly hazardous waste identification rules and land
       disposal restrictions, and for protecting the air, water, arid land by nunimizing and controlling
       all releases from fume hoods and bench operations. Compliance with all sewage discharge
       permits and regulations is also required.

15.2   Acids and samples at pH < 2 must be  neutralized before being disposed, or must be handled as
       hazardous waste.

15.3   For further information on waste management, consult The Waste Management Manual for
       Laboratory Personnel and Less is Better: Laboratory Chemical Management for Waste
       Reduction, both available  from the American Chemical Society's Department of Government
       Relations and Science Policy, 1155 16th Street NW, Washington, DC 20036.
 16.0  References

 16 1   Crecelius, E.A.; Bloom, N.S.; Cowan, C.E.; Jenne, E.A.. Speciation of Selenium and Arsenic in
        Natural Waters and Sediments, Volume 2: Arsenic Speciation. Final Report.  Prepared for
        Electric Power Research Institute, Palo Alto, CA by Battelle, Pacific Northwest Laboratories,
        Richland, WA. 1986.

 16.2   Andreae, M.O. 'Determination of Arsenic Species in Natural Waters,"  Anal Chem. 1977, 49,
        820.

 16.3   Method 1669, "Method for Sampling Ambient Water for Determination of Metals at EPA
        Ambient Criteria Levels," U.S. Environmental Protection Agency, Office of Water, Office of
        Science and Technology, Engineering and Analysis Division (4303), 401 M St SW,
        Washington, DC 20460 (January 1996).              J
 24
                                                                           Draft, January 1996

-------
                                                                                    Method 1632
  16 A    Patterson, C.C.; Settle, D.M. "Accuracy in Trace Analysis"; In National Bureau of Standards
         Special Publication 422; LaHeur, P.D., Ed., U.S. Government Printing Office, Washington
         DC, 1976.

  16.5    "Working with Carcinogens," Department of Health, Education, and Welfare, Public Health
         Service, Centers for Disease Control, NIOSH, Publication 77-206, Aug. 1977 NTIS PB-
         277256.

  16.6    "OSHA Safety and Health Standards, General Industry," OSHA 2206, 29 CFR 1910.

  16.7    "Safety in Academic Chemistry Laboratories," ACS Committee on Chemical Safety, 1979.

  16.8    "Standard Methods for the Examination of Water and Wastewater," 18th ed. and later
        revisions, American Public Health Association, 1015 15th Street NW, Washington DC 20005
         1-35: Section 1090 (Safety), 1992.
 17.0  Glossary

 The definitions and purposes below are specific to this method, but have been conformed to common
 usage as much as possible.

 17.1    Ambient Water—Waters in the natural environment (e.g., rivers, lakes, streams, and other
        receiving waters), as opposed to effluent discharges.

 17.2    Analytical Shift—All the 12-hour periods during which analyses are performed. The period
        begins with the purging of the OPR standard and ends exactly 12 hours later. All analyses
        both started and completed within this 12-hour period are valid.

 17.3    Calibration Standard (CAL)—A solution prepared from a dilute mixed standard and/or stock
        solutions and used to calibrate the response of the instrument with respect to analyte
        concentration.

 17.4    Equipment Blank—An aliquot of reagent water that is subjected in the laboratory to all
        aspects of sample collection and analysis, including contact with all sampling devices and
        apparatus.  The purpose of the equipment blank is to determine if the sampling devices and
        apparatus for sample collection have been adequately cleaned before shipment to the field site.
        An acceptable equipment blank must be achieved before the sampling devices and apparatus
        are used for sample collection. In addition, equipment blanks should be run on random,
        representative sets of gloves, storage bags, and plastic wrap for each lot to determine if these
        materials are free from contamination before use.
Draft, January 1996
                                                                                           25

-------
Method 1632
17 5   Field Blank—An aliquot of reagent water that is placed in a sample container in the
       laboratory, shipped to the field, and treated as a sample in all respects, including contact with
       the sampling devices and exposure to sampling site conditions, storage, preservation, and all
       analytical procedures, which may include filtration. The purpose of the field blank is to
       determine if the field or sample transporting procedures and environments have contaminated
       the sample.
                                                           I
17.6   Field Duplicates (FD1 and FD2>—Two separate samples collected in separate sample bottles
       at the same time and place under identical circumstances and treated exactly the same
       throughout field and laboratory procedures.  Analyses of FD1 and FD2 give a measure of the
       precision associated with sample collection, preservation, and storage, as weU as with
       laboratory procedures.

17 7   Initial Precision and Recovery (EPR)—Four aliquots of the ongoing precision and recovery
       standard analyzed to establish the ability to generate acceptable precision and accuracy.  IPRs
       are performed before a method is  used for the first time and any time the method or
       instrumentation is modified.

 17.8   Intercomparison Study—An exercise in which samples are prepared and split by a reference
       laboratory, then analyzed by one or more testing laboratories and the reference laboratory.
       The intercomparison, with a reputable laboratory as the reference laboratory, serves as the best
       test of the precision and accuracy of the analyses at natural environmental levels.

       Laboratory Blank—An aliquot of reagent water that is treated exactly as a sample including
        exposure to all glassware, equipment, solvents, reagents, internal standards, and surrogates that
        are used with samples.  The laboratory blank is used to determine if analytes or interferences
        are present in the laboratory environment, the reagents, or the apparatus.

 17.10  Laboratory Control Sample (LCS)—See Ongoing Precision and Recovery (OPR) Standard.

 17 11  Laboratory Duplicates (LD1 and LD2)—Two aliquots of the same sample taken in the
        laboratory from the same sample  bottle and analyzed separately using the referenced method.
        Analyses of LD1  and LD2 indicate precision associated with laboratory procedures, but not
        with sample collection, preservation, transportation, or storage procedures.

 17.12  Laboratory Fortified Blank—See Ongoing Precision and Recovery (OPR) Standard.

 17.13  Laboratory Fortified Sample Matrix—See Matrix Spike (MS) and Matrix Spike Duplicate
        (MSD).

 17.14  Laboratory Reagent Blank—See Laboratory Blank.

 17.15  Matrix Spike (MS) and Matrix Spike Duplicate (MSD)—:Aliquots of an environmental
        sample to which known quantities of the analytes are added in. the laboratory. The MS and
17.9
 26
                                                                              Draft, January 1996

-------
                                                                                      Method 1632
 17.16

 17.17

 17.18

 17.19


 17.20

 17.21
 17.22

 17.23


 17.24
 17.25


 17.26

 17.27
 MSD are analyzed exactly like a sample. Their purpose is to quantify the bias and precision
 caused by the sample matrix.  The background concentrations of the analytes hi the sample
 matrix must be determined hi a separate aliquot and the measured values in the MS and MSD
 corrected for background concentrations.

 May—This action, activity, or procedural step is optional.

 May Not—This action, activity, or procedural step is prohibited.

 Method Blank—See Laboratory  Blank.

 Minimum Level (ML)—The lowest level at which the entire analytical system gives a
 recognizable signal and acceptable calibration point.

 Must—This action, activity, or procedural step is required.

 Ongoing Precision and Recovery (OPR) Standard - A laboratory blank spiked with known
 quantities of analytes. The OPR is analyzed exactly like a sample.  Its purpose is to  assure
 that the results produced by the laboratory remain within the limits specified in the referenced
 methods for precision and accuracy.

 Preparation Blank—See Laboratory Blank.

 Primary Dilution Standard—A solution containing the analytes that is purchased or prepared
 from stock solutions and diluted as needed to prepare  calibration solutions and other solutions.

 Quality Control Sample (QCS)—A sample containing all or a subset of the analytes at
 known concentrations. The QCS  is obtained from a source external to the laboratory or is
 prepared from a source of standards different from the source of calibration standards. It is
 used to check laboratory performance with test materials prepared external to the normal
 preparation process.

 Reagent Water—Water demonstrated to be free from the metal(s) of interest and potentially
 interfering substances at the MDL for that metal hi the referenced method.

 Should—This action, activity, or procedural step is suggested but not required.

Stock Solution—A solution containing an analyte that is prepared using a reference material
traceable to EPA, the National Institute of Science and Technology (NIST),  or a source that
will attest to the purity and authenticity of the reference material.
Draft, January 1996
                                                                                             27

-------
Method 1632
                      TABLE 1: METHOD DETECTION LIMIT
                        (MDL) AND MINIMUM LEVEL (ML)
Method Detection Limit (MDL):
Minimum Level (ML):
0.002 pg/L
0.01000 ug/L
                         TABLE 2:  METHOD QC CRITERIA
IPR Average Recovery (X)
IPR Standard Deviations)
Matrix Spike Recovery (P)1
OPR Recovery (P)1
MS/MSD RPD
Calibration Verification
59-143%
< 42%
55-146%
55-1.46%
<20%
76-1.16%
                      1 OPS recoveiy required only SmaWx spike recoveiy does not meet oterti
 28
                                                                    Draft, January 1996

-------
                                                                Method 1632
                 TABLE 3: TYPICAL SPECTROPHOTOMETER
                                SETTINGS
Parameter
EDL energy
EDL power
Wavelength
Slit width
Typical Setting
59
8 W
193.7 nm
0.7 nm
     TABLE 4:  TYPICAL FLOWS AND PRESSURES FOR GASES IN THE HYDRIDE
                           GENERATION SYSTEM
          Gas
Flow Rate (mL/min)
Pressure (Ib/in2)
           He
           H2
           Air
      150
      350
      180
     10
     20
     20
Draft, January 1996
                                                                      29

-------
   Method 1632
   Figure 1.  Arsenic Speciation Apparatus:  (a) Quartz Cuvette Burner Tube, (b) Reaction Vessel, and

   (c) Schematic Diagram

                                                                1B
                                                                REACTION VESSEL
From Trap
     1C
     SCHEMATIC DIAGRAM
                                                                   ToVxuc
                                                                 Connector
                                                                 from Trap
                                                                 To Burner
                                                                  8O»0
                                                             Wire-wound
                                                             Pyrox U-Tub-s
                                                             Half-packed with
                                                             1 S% OV-3 oo
                                                             Chroma«xt> **W"
                                                                                Chart Recorder Output
    30
                                                                                              Draft, January 1996

-------