Method 1640: Determination of Trace
Elements in Ambient Waters by On-
Line Chelation Preconcentration and
Inductively Coupled Plasma-Mass
Spectrometry
                                   > Printed on Recycled Paper

-------
Method 1640
                                    Acknowledgments
Method 1640 was prepared under the direction of William A. Telliard  of the U.S. Environmental
Protection Agency's (EPA's) Office of Water (OW), Engineering and Analysis Division (BAD).  The
method was prepared under EPA Contract 68-C3-0337 by the DynCorp Environmental Programs Division
with assistance from Interface, Inc.

The  following researchers contributed to the philosophy behind this method.  Their contribution is
gratefully acknowledged:
                                                        l
Shier Berman, National Research Council, Ottawa, Ontario, Canada
Nicholas Bloom, Frontier Geosciences Inc, Seattle, Washington
Paul Boothe and Gary Steinmetz, Texas A&M University, College Station,  Texas
Eric Crecelius, Battelle Marine Sciences Laboratory, Sequim, Washington
Russell Flegal, University of California/Santa Cruz, California
Gary Gill, Texas A&M University  at Galveston, Texas
Carlton Hunt and Dion Lewis, Battelle Ocean Sciences, Duxbuiry, Massachusetts
Carl Watras, Wisconsin Department of Natural Resources, Boulder Junction, Wisconsin
Herb Windom and Ralph Smith, Skidaway Institute of Oceanography, Savannah, Georgia

Li addition, the following personnel at the EPA Office of Research and Development's Environmental
Monitoring Systems Laboratory in Cincinnati, Ohio, are gratefully acknowledged for the development of
the analytical procedures described in this method:

J.T. Creed
T.D. Martin
SJE. Long (DynCorp, formerly Technology Applications, Inc.)

                                                        i
                                        Disclaimer


This method has been reviewed and approved for publication by the Engineering and Analysis Division
of the U.S. Environmental Protection Agency. Mention of trade names or commercial products does not
constitute endorsement or recommendation for use.


Questions concerning this method or its application should be addressed to:

W.A. Telliard
USEPA Office of Water
Analytical Methods Staff
Mail Code 4303
401  M Street, SW
Washington, DC  20460
Phone: 202/260-7120
Fax:  202/260-7185
 K
January 1996

-------
                                                                                     Method 1640
 Introduction

 This analytical method was designed to support water quality monitoring programs authorized under the
 Clean Water Act.  Section 304(a) of the Clean Water Act requires EPA to publish water quality criteria
 that reflect the latest scientific knowledge concerning the physical fate (e.g., concentration and dispersal)
 of pollutants,  the effects of pollutants on ecological and human health, and the effect of pollutants on
 biological community diversity, productivity, and stability.

 Section 303 of the Clean Water Act requires states to set a water quality standard for each body of water
 within its boundaries.  A state water quality standard consists of a designated use or uses of a waterbody
 or a segment of a waterbody,  the water quality criteria that are necessary to protect the designated use or
 uses, and an antidegradation policy.  These water quality standards serve two purposes: (1) they establish
 the water quality goals for a specific waterbody, and (2) they are the basis for establishing water quality-
 based treatment controls and strategies beyond the technology-based controls required by Sections 301(b)
 and 306 of the Clean Water Act.

 In defining water quality standards,  the state may use narrative criteria,  numeric criteria, or both.
 However, the 1987 amendments to the Clean Water Act required states to adopt numeric criteria for toxic
 pollutants (designated in Section 307(a) of the  Act) based on EPA Section  304(a)  criteria or other
 scientific data, when the discharge or presence of those toxic pollutants could reasonably  be expected to
 interfere with  designated uses.

 In some cases, these water quality criteria are as much as 280 times lower than those that can be achieved
 using existing EPA methods and required to support technology-based permits. Therefore, EPA developed
 new sampling and analysis methods to specifically address state needs for measuring toxic metals at water
 quality criteria levels, when such measurements are necessary to protect designated uses in state  water
 quality standards.  The latest  criteria published by EPA are those listed in the National Toxics Rule (57
 FR 60848) and the Stay of Federal Water Quality  Criteria for Metals (60 FR 22228).  These rules include
 water quality  criteria for 13  metals, and it is  these criteria on which the new sampling and analysis
 methods are based. Method  1640 was specifically developed to provide reliable measurements of four
 of these metals at EPA WQC levels using on-line chelation preconcentration and inductively coupled
 plasma-mass spectrometry techniques.

 In developing  these methods, EPA found that one of the greatest difficulties in measuring pollutants at
 these levels was precluding sample contamination during collection, transport, and analysis. The degree
 of difficulty, however, is highly dependent on the metal and site-specific conditions.  This analytical
 method, therefore, is designed to provide the level of protection necessary to preclude contamination in
 nearly all situations.  It is also designed to provide the procedures necessary to produce reliable results
 at the lowest possible water quality criteria published by EPA.  In recognition of the variety of situations
 to which this method may be applied, and in recognition of continuing technological advances, the method
 is performance-based.  Alternative procedures may be used, so long as those procedures are demonstrated
 to yield reliable results.

 Requests for additional copies should be directed to:

 U.S. EPA NCEPI
 11029 Kenwood Road
 Cincinnati, OH 45242
 513/489-8190
January 1996

-------
Method 1640
    Note: This method is intended to be performance based, and the laboratory is permitted to omit any
    step or modify any procedure provided that all performance requirements set forth in this method
    are met The laboratory is not allowed to omit any quality control analyses. The terms "must,"
    "may,"  and "should" are included throughout this  method and  are intended to illustrate the
    importance of the procedures in producing verifiable data at water quality criteria levels. The term
    "must" is used to indicate that researchers in trace metals smalysis  have found certain procedures
    essential in successfully analyzing samples and avoiding contamination; however, these procedures
    can be modified or omitted if the laboratory can show that  data quality is not affected.
 IV
                                                                                    January 1996

-------
                                                                         Method 1640
                             Method  1640
   Determination of Trace Elements in Ambient  Waters  by
     On-Line  Chelation Preconcentration and Inductively
                Coupled  Plasma-Mass  Spectrometry
 1.0   Scope and Application

 1.1    This method is for the determination of dissolved elements in ambient waters at EPA water
       quality criteria (WQC) levels using on-line chelation preconcentration and inductively coupled
       plasma-mass spectrometry (ICP-MS). It may also be used for determination of total
       recoverable element concentrations in these waters. This method was developed by integrating
       the analytical procedures contained in EPA Method 200.10 with the quality control (QC) and
       sample handling procedures necessary to avoid contamination and ensure the validity of
       analytical results during sampling and analysis for metals at EPA WQC levels. This method
       contains QC procedures that will assure that contamination will be detected when blanks
       accompanying samples are analyzed. This method is accompanied by Method 1669:
       Sampling Ambient Water for Determination of Trace Metals at EPA Water Quality Criteria
       Levels (the "Sampling Method"). The Sampling Method is necessary to ensure that
       contamination will not compromise trace metals determinations  during the sampling process.

 1.2    This method is applicable to the following elements:
1.3
                                             Chemical Abstract
                 Analyte       Symbol       Services Registry
                                             Number (CASRN)
Cadmium
Copper
Lead
Nickel
(Cd)
(Cu)
(Pb)
(Ni)
7440-43-9
7440-50-8
7439-92-1
7440-02-0
Table 1 lists the EPA WQC levels, the method detection limit (MDL) for each metal, and the
minimum level (ML) for each metal in this method. Linear working ranges will be dependent
on the instrumentation and selected operating conditions but should be essentially independent
of the matrix because elimination of the matrix is a feature of the method.

This method is not intended for determination of metals at concentrations normally found in
treated and untreated discharges from industrial facilities. Existing regulations (40 CFR Parts
400-500) typically limit concentrations in industrial discharges to the mid to high part-per-
billion (ppb) range, whereas ambient metals concentrations  are normally in the low part-per-
trillion (ppt) to low ppb range.
January 1996

-------
Method 1640
1.4    The ease of contaminating ambient water samples with ithe metal(s) of interest and interfering
       substances cannot be overemphasized. This method includes suggestions for improvements in
       facilities and analytical techniques that should maximize the ability of the laboratory to make
       reliable trace metals determinations and minimize contamination.  These suggestions are given
       in Section 4.0 and are based on findings of researchers performing trace metals analyses
       (References 1-8).  Additional suggestions for improvement of existing facilities may be found
       in EPA's  Guidance for Establishing Trace Metals Clean Rooms in Existing Facilities, which is'
       available from the National Center for Environmental Publications and Information (NCEPI) at,
       the address listed in the introduction to this document.

1.5    Clean and ultraclean—The terms "clean" and "ultraclean" have been applied to the techniques  .
       needed to reduce or eliminate contamination  in trace metals  determinations. These terms are
       not used in this method because they lack an exact  definition.  However, the information
       provided in this method is consistent with the summary guidance  on clean and ultraclean      :
       techniques (Reference 9).

1.6    This method follows the EPA Environmental Methods Management Council's "Format for
       Method Documentation" (Reference 10).

1.7    This method is "performance based"; i.e., an  alternate procedure or technique may be used, as
       long as the performance requirements in the method are met. Section 9.1.2 gives details of the
       tests and documentation required to support and document equivalent performance.

1.8    For dissolved metal determinations, samples  must be filtered through a 0.45-um capsule filter
       at the field site.  The Sampling Method describes the filtering procedures.  The filtered
       samples may be preserved in the field or transported to  the laboratory for preservation.
       Procedures for field preservation are detailed in the Sampling Method; provides  procedures for
       laboratory preservation are provided in this method.

1.9    Acid solubilization is required before the determination of total recoverable elements to aid
       breakdown of complexes or colloids that might influence trace element recoveries.

1.10   This method should be used by analysts experienced in the use of inductively coupled plasma >
       mass spectrometry (TCP-MS), including the interpretation of spectral and matrix interferences
       and procedures for their correction; and should be used only by personnel thoroughly trained
       in the handling and analysis of samples for determination of metals at EPA WQC levels.  A
       minimum of six months' experience with commercial instrumentation is recommended.

1.11   This method is accompanied by a data verification  and validation guidance document,
       Guidance on the Documentation and Evaluation  of Trace Metals Data Collected for CWA
       Compliance Monitoring.  Before using this method, data users should state the data quality
       objectives (DQOs) required for a project.
                                                                                              '

2.0   Summary of Method

2.1    This method is used to preconcentrate trace elements using  an iminodiacetate functionalized
       chelating resin (References 11-12). Following acid solubilization, the sample is buffered prior
       to the chelating column using an on-line system. Group I and n  metals, as well as most
                                                                                   January 1996

-------
                                                                                    Method 1640
        onions, are selectively separated from the analytes by elution with ammonium acetate at pH
        5.5.  The analytes are subsequently eluted into a simplified matrix consisting of dilute nitric
        acid and are determined by ICP-rMS using a directly coupled on-line configuration.

 2.2    The determinative step in this method is ICP-MS (Reference 13-15).  Sample material in
        solution is introduced by pneumatic nebulization into a radiofrequency plasma where energy
        transfer processes cause desolvation, atomizatipn, and ionization. The ions are extracted from
        the plasma through a differentially pumped vacuum interface and separated on the basis of
        their mass-to-charge (m/z) ratio by a mass spectrometer having a minimum resolution
        capability of 1 amu peak width at 5% peak height at m/z 300.  An electron multiplier or
        Faraday detector detects ions transmitted through the mass analyzer, and a data handling
        system processes the  resulting current.  Interferences relating to the technique (Section 4) must
        be recognized and corrected.  Such corrections must include compensation for isobaric
        elemental interferences and interferences from polyatomic ions derived from the plasma gas,
        reagents, or sample matrix. Instrumental drift must be corrected for by the use of internal
        standardization.

 3.0    Definitions
 3.1
4.1
4.2
Apparatus—Throughout this method, the sample containers, sampling devices, instrumentation,
and all other materials and devices used in sample collection, sample processing, and sample
analysis activities will be referred to collectively as the Apparatus.
3.2     Other definitions of terms are given in the glossary (Section 18) at the end of this method.

4.0    Contamination and Interferences
Preventing ambient water samples from becoming contaminated during the sampling and
analytical process constitutes one of the greatest difficulties encountered in trace metals
determinations. Over the last two decades, marine chemists have come to recognize that much
of the historical data on the concentrations of dissolved trace metals in seawater are
erroneously high because the concentrations reflect contamination from sampling and analysis
rather than ambient levels. More recently, historical trace metals data collected from
freshwater rivers and streams have been shown to be similarly biased because of contamination
during sampling and analysis (Reference 16).  Therefore, it is imperative that extreme care be
taken to avoid contamination when collecting and analyzing ambient water samples for trace
metals.

There are numerous routes by which samples may become contaminated.  Potential sources of
trace metals contamination during sampling include metallic or metal-containing labware (e.g.,
talc gloves that contain high levels of zinc),  containers, sampling equipment, reagents, and
reagent water; improperly cleaned and stored equipment, labware, and reagents; and
atmospheric inputs such as dirt and dust. Even human contact can be a source of trace metals
contamination. For example, it has been demonstrated that dental work (e.g., mercury
amalgam fillings) in the mouths of laboratory personnel can contaminate samples that are
directly exposed to exhalation (Reference 3).
January 1996

-------
Method 1640
4.3    Contamination Control

       4.3.1   Philosophy—The philosophy behind contamination control is to ensure that any object
               or substance that contacts the sample is metal fnse and free from any material that may
               contain metals.

               4.3.1.1  The integrity of the results produced cannot be compromised by contamination
                      of samples. Requirements and suggestions for control of sample contamination
                      are given in this method and the Sampling Method.

               4.3.1.2  Substances in a sample cannot be allowed to contaminate the laboratory work
                      area or instrumentation used for trace metals measurements. Requirements and
                      suggestions for protecting the laboratory are given hi this method.

               4.3.1.3  Although contamination control is essential, personnel health and safety remain
                      the highest priority.  Requirements and suggestions for personnel safety are
                      given in Section 5 of this method and the Sampling Method.

    .   4.3.2   Avoiding contamination—The best way to control contamination is to completely
               avoid exposure of the sample to contamination in the first place. Avoiding exposure
               means performing operations in an area known to be free from contamination. Two of
               the most important factors hi avoiding/reducing sample contamination are (1) an
               awareness of potential sources of contamination and (2) strict attention to work being
               done. Therefore it is imperative that the procedures described hi this method be
               carried out by well-trained, experienced personnel.

       4.3.3   Use a clean environment—The ideal environment for processing samples is a class  100;
               clean room (Section 6.1.1).  If a clean room is not available, all sample preparation
               should  be performed hi a class 100 clean bench or a nonmetal glove box fed by
               particle-free air or nitrogen. Digestions should be performed in a nonmetal fume hood
               situated, ideally, hi the clean room.

       4.3.4   Minimize exposure—The Apparatus  that will contact samples, blanks, or standard
               solutions should be opened or exposed only in a clean room, clean bench, or glove
               box so that exposure to an uncontrolled atmosphere is minimized.  When not being
               used, the Apparatus should be covered with clean plastic wrap, stored in the  clean
               bench or in a plastic box or glove box, or bagged in clean zip-type bags. Minimizing
               the time between cleaning and use will also mhiimize contamination.

        4.3.5   Clean work surfaces—Before a given batch of samples is processed, all work surfaces
               in the hood, clean bench, or glove box hi which the samples will be processed should
               be cleaned by wiping with a lint-free cloth or wipe soaked with reagent water.

        4.3.6   Wear gloves—Sampling personnel must wear clean, nontalc gloves (Section  6.10.7)
               during all operations involving handling of the Apparatus, samples, and blanks.  Only
             • clean gloves may touch the Apparatus. If another object or substance is touched, the
               glove(s) must be changed before again handling the Apparatus.  If it is even suspected
               that gloves have become contaminated, work must be halted, the contaminated gloves
               removed, and a new pair of clean gloves put on.  Wearing multiple layers of clean
                                                                                   January 1996

-------
                                                                                     Method 1640
               gloves will allow the old pair to be quickly stripped with minimal disruption to the
               work activity.

        4.3.7  Use metal-free Apparatus—All Apparatus used for determination of metals at ambient
               water quality criteria levels must be nonmetallic, free of material that may contain
               metals, or both.

               4.3.7.1  Construction materials—Only the following materials should come in contact
                       with samples:  fluoropolymer (FEP, PTFE), conventional or linear
                       polyethylene, polycarbonate, polypropylene, polysulfone, or ultrapure quartz.
                       PTFE is less desirable than EEP because the sintered material in PTFE may
                       contain contaminates and is susceptible to serious memory contamination
                       (Reference 6).  Fluoropolymer or glass containers should be used for samples
                       that will be analyzed for mercury because mercury vapors can diffuse in or out
                       of the other materials resulting either in contamination or low-biased results
                       (Reference 3).  All materials, regardless of construction, that will directly or
                       indirectly contact the sample must be cleaned using the procedures described
                       hi Section 11 and must be known to be clean and metal-free before
                       proceeding.

               4.3.7.2  The following materials have been found to contain trace metals and should
                       not contact the sample or be  used to hold liquids that contact the sample,
                       unless these materials have been shown to be free of the metals of interest at
                       the desired level: Pyrex, Kimax, methacrylate, polyvinylchloride, nylon, and
                       Vycor (Reference 6). In addition, highly colored plastics, paper cap liners,  .
                       pigments used to mark increments on plastics, and rubber all contain trace
                       levels of metals and  must be avoided (Reference 17).

               4.3.7.3  Serialization—It is recommended that serial numbers be indelibly marked or
                       etched on each piece of Apparatus so that contamination can be traced, and
                       logbooks should be maintained to track the sample from the container through
                       the labware to injection into the instrument. It may be useful to dedicate
                       separate sets of labware to different sample types; e.g., receiving waters vs.
                       effluents.  However,  the Apparatus used for processing blanks and standards
                       must be mixed with the Apparatus used to process samples  so that
                       contamination of all labware can be detected.

               4.3.7.4  The laboratory  or cleaning facility is responsible for cleaning the Apparatus
                       used by the  sampling team.  If there are any indications that the Apparatus is
                      not clean when received by the sampling team (e.g., ripped  storage bags), an
                       assessment of the likelihood of contamination must be made.  Sampling must
                      not proceed if it is possible that the Apparatus is contaminated.  If the
                      Apparatus is contaminated, it must be returned to the laboratory or cleaning
                      facility for proper cleaning before any sampling activity resumes.

        4.3.8 " Avoid Sources of Contamination—Avoid contamination by being aware of potential
               sources  and routes of contamination.
January 1996

-------
Method 1640
               4.3.8.1 Contamination by carryover—Contamination may occur when a sample
                      containing low concentrations of metals is processed immediately after a
                      sample containing relatively high concentrations of these metals.  To reduce
                      carryover, the sample introduction system may be rinsed between samples with
                      dilute acid and reagent water.  When an unusually concentrated sample is
                      encountered, it is followed by analysis of a laboratory blank to check for      \
                      carryover. For samples containing high, levels of metals, it may be necessary
                      to acid-clean or replace the connecting tubing or inlet system to ensure that
                      contamination will not affect subsequent measurements. Samples known or
                      suspected to contain the lowest concentration of metals should be analyzed
                      first followed by samples containing hi|^ier levels. For instruments containing.
                      autosamplers, the laboratory should keep track of which station is used for a
                      given sample. When an unusually high concentration of a metal is detected in
                      a sample, the station used for that sample should be cleaned more thoroughly
                      to prevent contamination of subsequent samples, and the results for subsequent
                      samples should be checked for evidence of the metal(s) that occurred in high
                      concentration.

               4.3.8.2 Contamination by samples—Significant laboratory or instrument contamination
                      may result when untreated effluents, in-process waters, landfill leachates, and
                      other samples containing high concentrations of inorganic substances  are
                      processed and analyzed. As stated in Section 1.0, this method is not  intended
                      for application to these samples, and samples containing high concentrations
                      should not be permitted into the clean room and laboratory  dedicated for
                      processing trace metals samples.

               4.3.8.3 Contamination by indirect contact—Apjparatus that may not directly come in
                      contact with the samples may still be a source of contamination. For example,
                      clean tubing placed in a dirty plastic bag may pick up contamination  from the
                      bag and then subsequently transfer the contamination to the sample.
                      Therefore, it is imperative that every piece of the Apparatus that is directly or
                      indirectly used in the collection, processing, and analysis of ambient water
                      samples be  cleaned as specified in Section 11.

               4.3.8.4 Contamination by airborne particulate matter—Less obvious substances capable
                      of contaminating samples include airborne particles.  Samples may be
                      contaminated by airborne dust, dirt, paiticles, or vapors from:  unfiltered air
                      supplies; nearby corroded or rusted pipes, wires, or other fixtures; or metal-
                      containing paint. Whenever possible, sample processing and analysis should
                      occur as far as possible from sources of airborne contamination.

 4.4    Interferences—Interference  sources that may cause inaccuracies in the determination  of trace
        elements by ICP-MS are given below and must be recognized and corrected for. Internal
        standards should be used to correct for instrumental drift as well as suppressions or
        enhancements of instrument response caused by the sample matrix.

        4.4.1   Isobaric elemental interferences—Are caused by isotopes of different elements that
               form singly or doubly charged ions of the same nominal m/z and that cannot be
               resolved by the mass spectrometer.  All elements determined by this method have, at a
                                                                                    January 1996

-------
                                                                                     Method 1640
               minimum, one isotope free of isobaric elemental interferences. If an alternative
               isotope that has a higher natural abundance is selected to achieve greater sensitivity, an
               isobaric interference may occur. All data obtained under such conditions must be
               corrected by measuring the signal from another isotope of the interfering element and
               subtracting the contribution the isotope of interest based on the relative abundance of
               the alternate isotope and isotope of interest.  A record of this correction process should
               be included with the report of the data.  It should be noted that such corrections will
               only be as accurate as the accuracy of the relative abundance used in the equation for
               data calculations. Relative abundances should be established before any corrections are
               applied.

        4.4.2   Abundance sensitivity—Is a property defining the degree to which the  wings of a mass
               peak contribute to adjacent m/z's. Ion energy and quadruple operating pressure affect
               the abundance sensitivity.  Wing overlap interferences may result when a small m/z
               peak is being measured adjacent to a large one. The potential for these interferences
               should be recognized  and the spectrometer resolution adjusted to minimize them.

        4.4.3   Isobaric polyatomic ion interferences—Are caused by ions consisting of more than one
               atom which have the same nominal m/z as the isotope of interest, and which cannot be
               resolved by the mass spectrometer in use.  These ions are commonly formed in the
               plasma or interface system from support gases or sample components.  Such
               interferences must be  recognized,  and when they cannot be avoided by selecting
               alternative analytical isotopes, appropriate corrections must be made to the data.
               Equations for the correction of data should be established at the time of the  analytical
               run sequence because  the polyatomic ion interferences will be highly dependent on the
               sample matrix and chosen instrument conditions.

        4.4.4   Physical interferences—Are associated with the physical processes that govern the
               transport of sample into  the plasma, sample conversion processes hi the plasma, and
               the transmission of ions  through the plasma-mass spectrometer interface.  These
               interferences may result  in differences between instrument responses for the  sample
               and the calibration standards.  Physical  interferences may occur in the transfer of
               solution to the nebulizer (e.g., viscosity effects), at the point of aerosol formation  and
               transport to the plasma (e.g., surface tension), or during excitation and  ionization
               processes within the plasma itself. Internal standardization may be effectively used to
               compensate for many  physical interference effects (Reference 18). Internal standards
               ideally should have similar analytical behavior to the elements being determined.

        4.4.5   Memory interferences—Result when isotopes of elements in a previous sample
               contribute to the signals  measured in a new sample.  Memory  effects can result from
               sample deposition on the sampler  and skimmer cones, and from the buildup  of sample
               material in  the plasma torch and spray chamber.  The site where these effects occur
               depends on the element and can be minimized by flushing the system with a rinse
               blank between samples (Section 7.6.3).  The possibility of memory interferences
               should be recognized within an  analytical run and suitable rinse times should be used
               to reduce them.  The rinse times necessary for a particular element should be estimated
               before  it is  analyzed.  This estimation may be achieved by aspirating a standard
               containing elements corresponding to ten times the upper end of the linear range for a
               normal sample analysis period, followed by analysis of the rinse blank  at designated
January 1996

-------
Method 1640
               intervals.  The length of time required to reduce analyte signals below the ML should
               be noted.  Memory interferences may also be assessed within an analytical run by
               using a minimum of three replicate integrations for data acquisition.  If the integrated
               signal values drop consecutively, the analyst should be alerted to the possibility of a
               memory effect, and should examine the analyte concentration in the previous sample to
               identify if the memory effect was high. If a memory interference is suspected, the
               sample should be reanalyzed after a long rinse period.

       4.4.6   A principal advantage of this method is the  selfjctive elimination of species giving rise
               to polyatomic spectral interferences on certain transition metals  (e.g., removal of the
               chloride interference on vanadium).  As most of the sample matrix is removed, matrix-.
               induced physical interferences are also substantially reduced.                        ,

       4.4.7   Low recoveries may be encountered in the preconcentration cycle if the trace elements1
               are complexed by competing chelators in the ssunple or are present as colloidal
               material.  Acid solubilization pretreatment is used to improve analyte recovery and to
               minimize adsorption, hydrolysis, and precipitation effects.

5.0   Safety

5.1    The toxicity or carcinogenicity of reagents used hi this method have not been fully established.
       Each chemical should be regarded as a potential health hazard and exposure to these         '
       compounds should be as low as reasonably achievable.

       5.1.1   Each laboratory is responsible for maintaining a current awareness file of OSHA
               regulations for the safe handling of the chemicals specified in this method (References
               19-22).  A reference file of material safety data sheets (MSDSs) should also be
               available to all personnel involved in the chemical analysis.  It is also suggested that
               the laboratory perform personal hygiene monitoring of each analyst who uses this
               method and that the results of this monitoring l>e made available to the analyst. The
               references and bibliography at the end of Reference 22 are particularly comprehensive
               in dealing with the general subject of laboratory safety.

       5.1.2   Concentrated nitric and hydrochloric acids present various hazards and are moderately ,
               toxic and extremely irritating to skin and mucus membranes.  Use these reagents in a
               fume hood whenever possible and if eye or  skin contact occurs, flush with large
               volumes of water.  Always wear protective clothing and safety glasses or a shield for
               eye protection, and observe proper mixing when working with these reagents.

5.2    The acidification of samples containing reactive materials may result in the release of toxic
       gases such as cyanides or sulfides.  Samples should be acidified hi a fume hood.

5.3    All personnel handling environmental samples known to contain or to have been in contact
       with human waste should be immunized against known, disease-causative agents.

5.4    Analytical plasma sources emit radiofrequency radiation in addition to intense UV radiation.
       Suitable precautions should be taken to protect personnel from such hazards. The inductively
       coupled plasma should only be viewed with proper eye protection from UV emissions.
                                                                                    January 1996

-------
                                                                                   Method 1640
 6.0    Apparatus, Equipment, and Supplies
        Disclaimer:  The mention of trade names or commercial products in this method is for
        illustrative purposes only and does not constitute endorsement or recommendation for use by
        the Environmental Protection Agency. Equivalent performance may be achievable using  -
        apparatus and materials other than those suggested here.  The laboratory is responsible for
        demonstrating equivalent performance.


6.1     Facility

        6.1.1   Clean room—Class 100, 200-ft2 minimum, with down-flow, positive-pressure
               ventilation, air-lock entrances, and pass-through doors.

               6.1.1.1  Construction materials—-Nonmetallic, preferably plastic sheeting attached
                      without metal fasteners.  If painted, paints that do not contain the metal(s) of
                      interest should be used.

               6.1.1.2 Adhesive mats—for use at entry points to control dust and dirt from shoes.

        6.1.2   Fume hoods—nonmetallic, two minimum, with one installed internal to the clean
               room.

        6.1.3   Clean benches—class 100, one installed in the clean room;  the other adjacent to the
               analytical instrument(s) for preparation of samples and standards.

6.2     Preconcentration system—System containing no metal parts in the analyte flow path,
        configured as shown in Figure  1.

        NOTE: An alternate preconcentration system to the one described below may be used
       provided that all performance criteria listed in this method can be met. If low
        recoveries are encountered in the preconcentration cycle for a particular analyte, it
        may be necessary to use an alternate preconcentration system.

        6.2.1   Column—Macroporous iminodiacetate chelating resin (Dionex Metpac CC-1 or
               equivalent).

        6.2.2   Sample loop—10-mL loop constructed from narrow-bore, high-pressure inert tubing,
               Tefzel ETFE (ethylene tetra-fiuoroethylene) or equivalent.

        6.2.3   Eluent pumping system (PI)—Programmable-flow, high-pressure pumping system,
               capable of delivering either one of two eluents at a pressure up to 2000 psi and a flow
               rate of 1-5 mL/min.

        6.2.4   Auxiliary pumps
January 1996

-------
Method 1640
               6.2.4.1  On-line buffer pump (P2)—Piston pump (Dionex QIC pump or equivalent) for
                      delivering 2M ammonium acetate buffer' solution.

               6.2.4.2  Carrier pump (P3>—Peristaltic pump (Gilson Minipuls or equivalent) for
                      delivering 1% nitric acid carrier solution.

               6.2.4.3  Sample pump (P4)—Peristaltic pump for loading sample loop.

       6.2.5   Control valves—Inert, double-stack, pneumatically operated four-way slider valves
               with connectors.

       6.2.6   Argon gas supply regulated at 80-100 psi

       6.2.7   Solution reservoirs—Inert containers, e.g., high density polyethylene (HDPE), for
               holding eluent and carrier reagents.

       6.2.8   Tubing—High pressure, narrow bore, inert tubiitg (e.g., Tefzel ETFE or equivalent) for
               interconnection of pumps and valve assemblies and a minimum length for connection
               of the preconcentration system to the ICP-MS instrument.

6.3 Inductively coupled plasma mass spectrometer

       6.3.1   Instrument capable of scanning the mass range 5-250 amu with a minimum resolution
               capability  of 1-amu peak width at 5% peak heij^it.  Instrument may be fitted with a   '
               conventional or extended dynamic range detection system.

       6.3.2   Radio-frequency generator compliant with FCC regulations.

       6.3.3   Argon gas supply—High-purity grade (99.99%).  When analyses are conducted       i
               frequently, liquid argon is more economical andi requires less frequent replacement of
               tanks than compressed argon in conventional cylinders.

       6.3.4   A variable-speed peristaltic pump is required for solution delivery to the nebulizer.

       6.3.5   A mass-flow controller on the nebulizer gas supply is required. A water-cooled spray
               chamber may be of benefit hi reducing some types of interferences (e.g.,  from
               polyatomic oxide species).

       6.3.6   If an electron multiplier detector is being used, precautions should be taken, where
               necessary, to prevent exposure to high ion flux. Otherwise changes hi instrument
               response or damage to the multiplier may result Samples having high concentrations
               of elements beyond the linear range of the instrument and with isotopes falling within
               scanning windows should be diluted before analysis.

 6.4   Analytical balance—with capability to measure to  0.1 rag, for use hi weighing solids and for
       preparing standards.

 6.5   Temperature-adjustable hot plate—capable of maintaining a temperature of 95°C.
 10
January 1996

-------
                                                                                     Method 1640
 6.6     Centrifuge with guard bowl, electric timer, and brake (optional)

 6.7     Drying oven—gravity convection, with thermostatic control capable of maintaining 105°C (±
        5°C).

 6.8     Alkaline detergent—Liquinox®, Alconox®, or equivalent.

 6.9     pH meter or pH paper

 6.10    Labware—For determination of trace levels of elements, contamination and loss are of prime
        consideration.  Potential contamination sources include improperly cleaned laboratory
        apparatus and general contamination within the laboratory environment from dust, etc.  A
        clean laboratory work area should be designated for handling trace element samples.  Sample
        containers can introduce positive and negative errors in the determination of trace elements by
        (1) contributing contaminants through surface desorption or leaching, and (2) depleting element
        concentrations through adsorption processes. All labware must be metal free.  Suitable
        construction materials are fluoropolymer (FEP, PTFE), conventional or linear polyethylene,
        polycarbonate, and polypropylene.  Fluoropolymer should be used when samples are to be
        analyzed for mercury. All labware should be cleaned according to the procedure hi Section
        11.4.  Gloves, plastic wrap, storage bags, and filters may all be used new without additional
        cleaning unless results of the equipment blank pinpoint any of these materials as a source of
        contamination. In this case, either an alternate supplier must be obtained or the materials must
        be cleaned.
        NOTE: Chromic acid must not be used for cleaning glassware.

        6.10.1  Volumetric flasks, graduated cylinders, funnels, and centrifuge tubes

        6.10.2  Assorted calibrated pipets

        6.10.3  Beakers—fluoropolymer (or other suitable material), 250-mL with fluoropolymer
               covers.

        6.10.4  Storage bottles—Narrow-mouth, fluoropolymer with fluoropolymer screw closure, 125-
               to 250-mL capacities.

        6.10.5  Wash bottle—One-piece stem fluoropolymer, with screw closure, 125-mL capacity.

        6.10.6  Tongs—For removal of Apparatus from acid baths.  Coated metal tongs may not be
               used.

        6.10.7  Gloves—clean, nontalc polyethylene, latex, or vinyl; various lengths.  Heavy gloves
               should be worn when working in acid baths since baths will contain hot, strong acids.

        6.10.8  Buckets or basins—5- to 50-L capacity, for acid soaking of the Apparatus.

        6.10.9  Brushes—Nonmetallic, for scrubbing Apparatus.
January 1996
11

-------
Method 1640
       6.10.10        Storage bags—dean, zip-type, nonvented, colorless polyethylene (various
                      sizes) for storage of Apparatus.

       6.10.11        Plastic wrap—Clean, colorless polyethylene for storage of Apparatus.
                                                                                                i
6.11   Sampling Equipment—The sampling team may contract with the laboratory or a cleaning
       facility that is responsible for cleaning, storing, and shipping all sampling devices, sample -    ,
       bottles, filtration equipment, and all other Apparatus used for the collection of ambient water
       samples.  Before the equipment is shipped to the field site, the laboratory or facility must
       generate an acceptable equipment blank (Section 9.6.3) to demonstrate that the sampling
       equipment is free from contamination.

       6.11.1  Sampling Devices—Before ambient water samples  are collected,  consideration should
               be given to the type of sample to be collected and the devices to  be used (grab,
               surface, or subsurface samplers).  The laboratory or cleaning facility must clean all
               devices used for sample collection.  Various types of samplers are described in the
               Sampling Method. Cleaned sampling devices should be stored in polyethylene bags or
               wrap.

       6.11.2  Sample bottles—Fluoropolymer, conventional or linear polyethylene, polycarbonate, or ;
               polypropylene; 500-mL with lids. Cleaned sample bottles should be filled with 0.1%
               HC1 (v/v) until use.                                                               \
        NOTE: If mercury is a target analyte, fluoropolymer or glass bottles must be used.	

        6.11.3  Filtration Apparatus

               6.11.3.1        Filter—Gelman Supor 0.45-um, 15-mm diameter capsule filter
                              (Gelman 12175, or equivalent).

               6.11.3.2        Peristaltic pump—115-V a.c., 12-V d.c., internal battery, variable
                              speed, single-head (Cole-Parmer, portable, "Masterflex L/S," Catalog
                              No. H-07570-10 drive with Quick Load pump head, Catalog No. H-
                              07021-24, or equivalent).

               6.11.3.3        Tubing for use with peristaltic pump—styrene/ethylene/butylene/
                              silicone (SEES) resin, approx 3/8-in i.d. by approximately 3 ft (Cole-
                              Parmer size 18, Catalog No. G-06464-18, or approximately  1/4-in i.d.,
                              Cole-Parmer size 17, Catalog No. G-06464-17, or equivalent).  Tubing
                              is cleaned by soaking in 5-10% HC1 solution for 8-24 h, rinsing with
                              reagent water in a clean bench in a clean room, and drying  in the  clean
                              bench by purging with metal-free air or nitrogen. After drying, the
                              tubing is double-bagged in clear polyethylene bags, serialized with a
                              unique number, and stored until use.
 12
January 1996

-------
                                                                                   Method 1640
 7.0    Reagents and Standards

        Reagents may contain elemental impurities that might affect the integrity of analytical data.
        Because of the high sensitivity of ICP-MS, high-purity reagents should be used. Each reagent
        lot should be tested for the metals of interest by diluting and analyzing an aliquot from the lot
        using the techniques and instrumentation to be used for analysis of samples.  The lot will Be
        acceptable if the concentration of the metal of interest is below the MDL listed in this method.
        All acids used for this method must be ultra high-purity grade.  Suitable acids are available
        from a number of manufacturers or may be prepared by sub-boiling distillation.

 7.1     Reagents for cleaning Apparatus, sample bottle storage, and sample preservation and analysis

        7.1.1   Nitric acid—concentrated (sp gr 1.41), Seastar or equivalent

        7.1.2   Nitric acid (1+1)—Add 500 mL concentrated nitric acid to 400 mL of regent water
               and dilute to 1 L.

        7.1.3   Nitric acid (1+9)—Add 100 mL concentrated nitric acid to 400 mL of reagent water
               and dilute to 1 L.

        7.1.4   Nitric acid 1.25M—Dilute 79 mL (112 g) concentrated nitric acid to 1000 mL with
               reagent water.

        7.1.5   Nitric acid 1%—Dilute  10 mL concentrated nitric acid to 1000 mL with reagent water.

        7.1.6   Hydrochloric acid—concentrated (sp gr 1.19).

        7.1.7   Hydrochloric acid (1+1)—Add 500 mL concentrated hydrochloric acid to 400 mL of
               reagent water and dilute to 1 L.

        7.1.8   Hydrochloric acid (1+4)—Add 200 mL concentrated hydrochloric acid to 400 mL of
               reagent water and dilute to 1 L.

        7.1.9   Hydrochloric acid (HO)—IN trace metal grade

        7.1.10  Hydrochloric acid (HC1)—10% wt, trace metal  grade

        7.1.11   Hydrochloric acid (HC1)—1% wt, trace metal grade

        7.1.12  Hydrochloric acid (HC1)—0.5% (v/v), trace metal grade

        7.1.13   Hydrochloric acid (HC1)—0.1% (v/v) ultrapure  grade

        7.1.14  Acetic acid, glacial  (sp gr 1.05)

        7.1.15   Ammonium hydroxide (20%)
January 1996
                                                                                           13

-------
Method 1640
       7.1.16  Ammonium acetate buffer 1M, pH 5.5—Add 58 mL (60.5 g) of glacial acetic acid to
               600 mL of reagent water.  Add 65 mL (60 g) of 20% ammonium hydroxide and mix.
               Check the pH of the resulting solution by withdrawing a small aliquot and testing with
               a calibrated pH meter, adjusting the solution to pH 5.5 (± 0.1) with small volumes of
               acetic acid or ammonium hydroxide as necessary.  Cool and dilute to 1 L with reagent
               water.

       7.1.17  Ammonium acetate buffer 2M, pH 5.5—Prepare as for Section 7.1.16 using 116 mL
               (121 g) glacial acetic acid and 130 mL (120 g) 20%  ammonium hydroxide, diluted to
               1000 mL with reagent water.
       NOTE: The ammonium acetate buffer solutions may be further purified by passing
       them through the chelating column at a flow rate of 5.0 mLlmin. With reference to
       Figure 1, pump the buffer solution through the column using pump PI, -with valves A
       and B off and valve C on.  Collect the purified solution in a container at the waste
       outlet. Then elute the collected contaminants from the column using 1.25M nitric acid
       for 5 min at a flow rate of 4.0 mLlmin.	

       7.1.18  Oxalic acid dihydrate (CASRN 6153-56-6), 0.2M—Dissolve 25.2 g reagent grade
               C2H2O4.2H2O hi 250 mL reagent water and dilute to 1000 mL with reagent water.
        CAUTION: Oxalic acid is toxic; handle with care.
7.2    Reagent water—Water demonstrated to be free from the metal(s) of interest and potentially
       interfering substances at the MDL for that metal listed in Table 1.  Prepared by distillation,
       deionization, reverse osmosis, anodic/cathodic stripping voltammetry, or other technique that
       removes the metal(s) and potential interferent(s).

7.3    Standard stock solutions—May be purchased from a reputable commercial source or prepared
       from ultra high-purity grade chemicals or metals (99.99-99.999% pure). All salts should be
       dried for 1 h at 105°C, unless otherwise specified.
        CAUTION: Many metal salts are extremely toxic if inhaled or swallowed. (Wash
        hands thoroughly after handling.) Stock solutions should be stored in plastic bottles.

        The following procedures may be used for preparing standard stock solutions:
        NOTE: Some metals, particularly those that form surface oxides, require cleaning
        before they are weighed.  This may be achieved by pickling the surface of the metal in
        acid. 'An amount over the desired weight should be pickled repeatedly, rinsed with
        water, dried, and weighed until the desired weight is achieved.
 14
January 1996

-------
                                                                                    Method 1640
        7.3.1   Bismuth solution, stock 1 mL = 1000 ug Bi—Dissolve 0.1115 g Bi2O3 in 5 mL
               concentrated nitric acid. Heat to effect solution.  Cool and dilute to 100 mL with
               reagent water.

        7.3.2   Cadmium solution, stock 1 mL = 1000 ug Cd: Pickle cadmium metal in (1+9) nitric
               acid to an exact weight of 0.100 g. Dissolve in 5 mL (1+1) nitric acid, heating to
               effect solution.  Cool and dilute to  100 mL with reagent water.

        7.3.3   Copper solution, stock  1 mL = 1000 ug Cu: Pickle copper metal in (1+9) nitric acid to
               an exact weight of 0.100 g. Dissolve hi 5 mL (1+1) nitric acid, heating to effect
               solution. Cool  and dilute to 100 mL with reagent water.

        7.3.4   Indium solution, stock  1 mL = 1000 ug In: Pickle indium metal in (1+1) nitric acid to
               an exact weight of 0.100 g.  Dissolve in 10 mL (1+1) nitric acid, heating to effect
               solution. Cool  and dilute to 100 mL with reagent water.

        7.3.5   Lead solution, stock 1 mL = 1000 ug Pb: Dissolve 0.1599 g PbNO3 in 5 mL (1+1)
               nitric acid. Dilute to 100 mL with reagent water.

        7.3.6   Nickel solution, stock 1 mL = 1000 ug Ni: Dissolve 0.100 g nickel powder in 5 mL
               concentrated nitric acid, heating to effect solution. Cool and dilute to 100 mL with
               reagent water.

        7.3.7   Scandium solution, stock 1 mL = 1000 ug Sc: Dissolve 0.1534 g Sc2O3 in 5 mL (1+1)
               nitric acid, heating to effect solution.  Cool and dilute to 100 mL with reagent water.

        7.3.8   Terbium solution, stock 1 mL = 1000 ug Tb: Dissolve 0.1176 g Tb4O7 in  5 mL
               concentrated nitric acid, heating to effect solution. Cool and dilute to 100 mL with
               reagent water.

        7.3.9   Yttrium solution, stock 1 mL = 1000 ug Y—Dissolve 0.1270 g Y2O3 in 5 mL (1+1)
               nitric acid, heating to effect solution.  Cool and dilute to 100 mL with reagent water.

7.4     Multielement stock standard solution—When multielement stock standards are prepared, care
        must be taken that the elements are compatible and stable. Originating element stocks should
        be checked for the presence of  impurities that might influence the accuracy of the standard.
        Freshly prepared standards should be transferred to acid-cleaned, new FEP or HDPE bottles
        for storage and monitored periodically for stability. A multielement stock standard solution
        containing cadmium, copper, lead, and nickel (1 mL = 10 ug) may be prepared by diluting 1
        mL of each single element stock in the list  to 100 mL with reagent water containing 1% (v/v)
        nitric acid.

        7.4.1   Preparation of calibration standards—Fresh multielement calibration standards should
               be prepared every 2 weeks or as needed.  Dilute the stock multielement standard
               solution to levels appropriate to the operating range of the instrument using reagent
            •  water containing 1% (v/v) nitric acid.  Calibration standards should be prepared at a
               minimum of three concentrations, one of which must be at the ML (Table 1), and
               another that must be near the upper end of the linear dynamic range.  If the direct
               addition procedure is being used (Method A, Section 10.3), add internal standards
January 1996
15

-------
Method 1640
               (Section 7.5) to the calibration standards and store in fluoropolymer bottles.
               Calibration standards should be verified initially using a quality control sample
               (Section 7.8).

7.5    Internal standard stock solution—1 mL = 100 ug.  Dilute 10 mL of scandium, yttrium, indium,
       terbium and bismuth stock standards (Section 7.3) to 100 mL with reagent water, and store in
       a FEP bottle.  Use this solution concentrate for addition to blanks, calibration standards and
       samples, or dilute by an appropriate amount using 1% (v/v) nitric acid, if the internal standards
       are being added by peristaltic pump (Method B, Section 10.3).
        NOTE: Bismuth should not be used as an internal standard using the direct addition
        method (Method A, Section 10.3) because it is not efficiently concentrated on the
        iminodiacetate column.          	__^_	

7.6     Blanks—The laboratory should prepare the following types of blanks. A calibration blank is
        used to establish the analytical calibration curve; the laboratory (method) blank is used to
        assess possible contamination from the sample preparation procedure and to assess spectral
        background; and the rinse blank is used to flush the instrument between samples in order to
        reduce memory interferences. In addition to these blanks, the laboratory may be required to   ;
        analyze field blanks (Section 9.6.2) and equipment blanks (Section 9.6.3).

        7.6.1   Calibration blank—Consists of 1% (v/v) nitric  acid in reagent water.  If the direct
               addition procedure  (Method A, Section 10.3) is being used, add internal standards.

        7.6.2   Laboratory blank—Must contain all the reagents in the same volumes as used in
               processing the samples. The laboratory blank must be carried through the same entire
               preparation scheme as the samples including digestion, when applicable (Section
               9.6.1).  If the direct addition procedure (Method A, Section 10.3) is being used, add
               internal standards to the solution after preparation is complete.

        7.6.3   Rinse blank—Consists of 1% (v/v) nitric acid in. reagent water.

7.7     Tuning solution—This solution is used for instrument tuning and mass calibration before
        analysis (Section 10.2). The solution is prepared by mixing nickel, yttrium, indium, terbium,
        and lead stock solutions (Section 7.3) in 1% (v/v) nitric acid to produce a concentration of 100
        ug/L of each element. Internal standards are not added to this solution.  (Depending on the
        sensitivity of the instrument, this solution may need to be diluted 10-fold.)

7.8     Quality control sample (QCS)—The QCS  should be obtained from a source outside the
        laboratory.  The concentration of the QCS solution analyzed will depend on the sensitivity of
        the instrument To prepare the QCS, dilute an appropriate aliquot of analytes to a concentration
        < 100  ug/L in 1% (v/v) nitric acid.  Because of lower  sensitivity, selenium may be diluted to a
        concentration of < 500 ug/L. If the direct addition procedure (Method A, Section 10.3) is
        being used, add internal standards after dilution,, mix, and store in a FEP bottle.   The QCS
        should be analyzed as needed to meet data quality needs and a fresh solution should be
        prepared quarterly or more frequently as needed.
 16
January 1996

-------
                                                                                   Method 1640
7.9    Ongoing precision and recovery (OPR) Sample—To an aliquot of reagent water, add aliquots
       of the multielement stock standard (Section 7.4) to prepare the OPR. The OPR must be
       carried through the same entire preparation scheme  as the samples including sample digestion,
       when applicable (Section 9.7). If the direct addition procedure (Method A, Section 10.3) is
       being used, add internal standards to this solution after preparation has been completed.


8.0   Sample Collection,  Filtration, Preservation, and Storage

8.1    Before an aqueous sample is collected, consideration should be given to the type of data
       required, (i.e., dissolved or total recoverable), so that appropriate preservation and pretreatment
       steps can be taken. The pH  of all aqueous samples must be tested immediately before they are
       aliquotted for processing or direct analysis to ensure the sample has  been properly preserved.
       If properly acid-preserved, the sample can be held up to 6 months before analysis.

8.2    Sample collection—Samples are collected as described in the Sampling Method.

8.3    Sample filtration—For dissolved metals, samples and field blanks are filtered through a 0.45-
       um capsule filter at the field site.  Filtering procedures are described in the Sampling Method.
       For the determination of total recoverable elements, samples are not filtered but should be
       preserved according to the procedures in Section 8.4.

8.4    Sample preservation—Preservation of samples and field blanks for both dissolved and total
       recoverable elements may be performed in the field when the samples are collected or in the
       laboratory. However, to avoid the hazards of strong acids in the field and transport
       restrictions, to minimize the potential for sample contamination, and to expedite field
       operations, the sampling team may prefer  to ship the samples to the  laboratory within 2 weeks
       of collection.  Samples and field blanks should be preserved at the laboratory immediately
       when they are received. For all metals, preservation involves the addition of 10% HNO3
       (Section 7.1.3) to bring the sample to pH <2.  For samples received  at neutral pH, approx 5
       mL of 10% HNO3  per liter will be required.

       8.4.1  Wearing clean gloves, remove the cap from the sample bottle, add die volume of
              reagent grade acid that will bring the pH to  <2, and recap the bottle immediately.  If
              the bottle is full, withdraw the necessary volume using a precleaned pipet and then add
              the acid. Record the volume  withdrawn and the amount of acid used.

       NOTE: Do not dip pH paper or a pH meter into the sample; remove a small aliquot
       •with a clean pipet  and test the aliquot.  When  the nature  of the sample is either
       unknown or known to be hazardous, the sample should be acidified in a fume hood.
       See Section 5.2.

       8.4.2  Store the preserved sample for a minimum of 48 h at 0-4°C to allow the acid to
              completely dissolve the metal(s) adsorbed on the container walls.  The sample pH
            . should be verified as <2 immediately before an aliquot is withdrawn for processing or
              direct analysis. If, for some reason such as  high alkalinity, the sample pH is verified
              to be >2, more acid must be added and the sample held for 16 h until verified to be
              pH<2. See Section  8.1.
January 1996
17

-------
Method 1640
       8.4.3  With each sample batch, preserve a method blank and an OPR sample hi the same way
              as the sample(s).

     .  8.4.4  Sample bottles should be stored hi polyethylene bags at 0-4°C until analysis.


9.0   Quality Assurance/Quality Control

9.1    Each laboratory that uses this method is required to operate a formal quality assurance
       program (Reference 23). The minimum requirements of this program consist of an initial
       demonstration of laboratory capability, analysis of samples spiked with metals of interest to
       evaluate and document data quality, and analysis of standards and blanks as tests of continued
       performance.  Laboratory performance is compared to established performance criteria to
       determine that results of the analysis meet the performance characteristics of the method.

       9.1.1  The analyst shall make an initial demonstration of the ability to generate acceptable
              accuracy and precision with this method. This ability is established as described in
              Section 9.2.                                                                      ;

       9.1.2  In recognition of advances that are occurring in analytical technology, the analyst is
              permitted to exercise  certain options to eliminate interferences or lower the costs of
              measurements.  These options Include alternate digestion, preconcentration, cleanup
              procedures, and changes in instrumentation.  Alternate determinative techniques, such
              as the substitution of  a colorimetric technique or changes that degrade method         ;
              performance, are not allowed. If an analytical technique other than the techniques
              specified hi the method is used, then that technique must have a specificity equal to or
              better than the specificity of the techniques in the method for the analytes of interest.

              9.1.2.1 Each time the method is modified, the analyst is required to repeat the
                      procedure in Section 9.2.   If the the change will affect the  detection limit of
                      the method, the laboratory is required to demonstrate that the MDL (40 CFR    ',
                      Part 136, Appendix  B) is lower than the MDL for that analyte in this method,
                      or one-third the regulatory compliance level, whichever is higher.  If the
                      change will affect calibration, the analyst must recalibrate the instrument
                      according to Section 10.

               9.1.2.2 The laboratory is required to maintain records of modifications made to  this
                      method. These records include the following, at a minimum:

                      9.1.2.2.1       The names, titles, addresses, and telephone numbers of the
                                     analyst(s) who performed the analyses and modification, and of
                                     the quality control officer who witnessed and will verify the
                                     analyses and modification.

                      9.1.2.2.2       A listing of metals measured, by name and CAS Registry
                                     number.

                      9.1.2.2.3       A narrative stating reason(s) for the modification(s).
18
January 1996

-------
                                                                                      Method 1640
                       9.1.2.2.4
                       9.1.2.2.5
                              Results from aULquality control (QC) tests comparing the
                              modified method to this method, including:

                              (a)     Calibration
                              (b)     Calibration verification
                              (c)     Initial precision and recovery (Section 9.2)
                              (d)     Analysis of blanks
                              (e)     Accuracy assessment

                              Data that will allow an independent reviewer to validate each
                              determination by tracing the instrument output (peak height,
                              area, or other signal) to the final result. These data are to
                              include, where possible:

                              (a)     Sample numbers and other identifiers
                              (b)     Digestion/preparation or extraction dates
                              (c)     Analysis dates and times
                              (d)     Analysis sequence/run chronology
                              (e)     Sample weight or volume
                              (f)      Volume before the extraction/concentration step
                              (g)     Volume after each extraction/concentration step
                              (h)    Final volume before analysis
                              (i)     Injection volume
                              (j)     Dilution data, differentiating between dilution of a
                                      sample or extract
                              (k)    Instrument and operating  conditions  (make, model,
                                     revision, modifications)
                              (1)     Sample introduction system (ultrasonic nebulizer, flow
                                     injection system, etc.)
                              (m)    Preconcentration system
                              (n)    Operating conditions (background corrections,
                                     temperature program, flow rates, etc.)
                              (o)    Detector (type, operating conditions, etc.)
                              (p)    Mass spectra, printer tapes,  and other recordings of raw
                                     data
                              (q)    Quantitation reports, data system outputs, and other
                                     data to link raw data to results reported

9.1.3   Analyses of blanks are required to demonstrate freedom from contamination.  Section
       9.6 describes the required types, procedures, and criteria for analysis of blanks.

9.1.4   The laboratory shall spike  at least 10% of the samples with the metal(s) of interest to
       monitor method performance.  Section 9.3 describes this test.  When results of these
       spikes indicate atypical method performance for samples, an alternative extraction or
       cleanup technique must be used to bring method performance within acceptable limits.
       If method performance for spikes cannot be brought within the limits given in this
       method, the result may not be reported for regulatory compliance purposes.
January 1996
                                                                                       19

-------
Method 1640
        9.1.5   The laboratory shall, on an ongoing basis, demonstrate through calibration verification
               and through analysis of the ongoing precision and recovery aliquot that the analytical
               system is hi control. These procedures are described in Sections 10.5 and 9.7 of this
               method.

        9.1.6   The laboratory shall maintain records to define title quality of data that are generated.
               Development of accuracy statements is described in  Section 9.3.4.

9.2     initial demonstration of laboratory capability

        9.2.1   Method detection limit—To establish the ability to detect the trace metals of interest,
               the analyst shall determine the MDL for each analyte according to the procedure in 40
               CFR 136, Appendix B using the apparatus, reagents, and standards that will be used in
               the practice of this method. The laboratory must produce an MDL that is less than or
               equal to the MDL listed in Table 1, or one-third the regulatory compliance limit,
               whichever is greater.  MDLs should be determinf jd when a new operator begins work
               or whenever, in the judgment of the analyst, a change in instrument hardware or
               operating conditions would dictate that they be redetermined.

        9.2.2   Initial precision and recovery (IPR)—To establish the ability to  generate acceptable
               precision and recovery, the analyst shall perform the following operations.

               9.2.2.1 Analyze four aliquots of reagent water spiked with the metal(s) of interest at
                      2-3 times the ML (Table 1), according to the procedures in Section 12. All
                      digestion, extraction, and concentration steps, and the containers, labware, and
                      reagents that will be used with samples must be used in this test.

               9.2.2.2 Using results of the set of four analyses, compute the average percent recovery
                      (X) for the metal(s) in each aliquot and the standard deviation of the recovery
                      (s) for each metal.

               9.2.2.3 For each metal, compare s and X with the corresponding limits for initial
                      precision  and recovery in Table 2. If s and  X for all metal(s) meet the
                      acceptance criteria, system performance is acceptable and analysis of blanks
                      and samples may begin. If, however, any individual s exceeds the precision
                      limit or any individual X falls outside the range for accuracy, system
                      performance is unacceptable for that metal.  Correct the problem  and repeat the
                      test (Section 9.2.2.1).

        9.2.3   Linear calibration ranges-r-Linear calibration ranges are primarily detector limited.
               The upper limit of the linear calibration range should be established for each analyte
               by determining the  signal responses from a minimum of three different concentration
               standards, one of which is close to the upper limit of the linear range.  Care should be
               taken to avoid potential damage to the detector during this process. The  analyst
               should judge the  linear calibration range that may be used for the analysis of samples
              " from the resulting data. The upper limit should be an observed signal no  more than
               10%  below the level extrapolated from lower standards.  Determined sample analyte
               concentrations that  are greater than  90% of the determined upper limit must be diluted
               and reanalyzed. The upper limits should be verified whenever, in the judgement of the
 20
January 1996

-------
                                                                                   Method 1640
               analyst, a change in analytical performance caused by either a change in instrument
               hardware or operating conditions would dictate they be redetermined.

       9.2.4   Quality control sample (QCS)—When beginning the use of this method, quarterly or
               as required to meet data quality needs, verify the calibration standards and acceptable
               instrument performance with the preparation and analyses of a QCS (Section 7.8).  To
               verify the calibration  standards the determined mean concentration from 3 analyses'of
               the QCS must be within ± 10% of the stated QCS value.  If the QCS is not within the
               required limits, an immediate second analysis of the QCS is recommended to confirm
               unacceptable performance. If the calibration standards, acceptable instrument
               performance, or both  cannot be verified, the source of the problem must be identified
               and corrected before proceeding with further analyses.

9.3    Method accuracy—To assess  the performance of the method on a given sample matrix, the
       laboratory must perform matrix spike (MS) and matrix spike duplicate (MSD) sample analyses
       on 10% of the samples from each site being monitored, or at least one MS sample analysis
       and one MSD sample analysis must be performed for each sample batch (samples collected
       from the same site at the same time,  to a maximum of 10 samples), whichever is  more
       frequent. Blanks (e.g., field blanks) may not be used for MS/MSD analysis.

       9.3.1   The concentration of the MS and MSD is determined as follows:

               9.3.1.1 If, as in compliance monitoring, the concentration of a specific metal in the
                      sample is being checked against a regulatory concentration limit, the spike
                      must be at that limit or at 1-5 times the background concentration, whichever
                      is greater.

               9.3.1.2 If the concentration is not being checked against a regulatory limit, the
                      concentration  must be at 1-5 times  the background concentration  or at 1—5
                      times the ML in Table 1, whichever is greater.

       9.3.2   Assessing spike recovery

               9.3.2.1 Determine the background concentration (B) of each metal by analyzing one
                      sample aliquot according to the procedure in Section 12.

               9.3.2.2 If necessary, prepare  a QC check sample concentrate that will produce the
                      appropriate level (Section 9.3.1) in the sample when the concentrate is added.

               9.3.2.3 Spike a second sample aliquot with the QC check sample concentrate and
                      analyze it to determine the concentration after spiking (A) of each metal.

               9.3.2.4 Calculate each percent recovery (P) as 100(A - B)/T, where T is the known
                      true value of the spike.

       9.3.3 •  Compare the percent recovery (P) for each metal with the corresponding QC
               acceptance criteria found in Table 2.  If any individual P falls outside the designated
               range for recovery, that metal has failed the acceptance criteria.
January 1996
21

-------
Method 1640
               9.3.3.1 For a metal that has failed the acceptance criteria, analyze the ongoing
                      precision and recovery standard (Section  9.7).  If the OPR is within its
                      respective limit for the metal(s) that failed (Table 2), the analytical system is in
                      control and the problem can be attributed to the sample matrix.

               9.3.3.2 For samples that exhibit matrix problems, furfher isolate the metal(s) from the
                      sample matrix using dilution, chelation, extraction, concentration, hydride ~
                      generation, or other means, and repeat the accuracy test (Section 9.3.2).

               9.3.3.3 If the recovery for the metal remains outside the acceptance criteria, the
                      analytical result for that metal in the unspiked sample is suspect and may not
                      be reported for regulatory compliance purposes.

        9.3.4   Recovery for samples should be assessed and records maintained.                     I

               9.3.4.1 After the analysis of five samples of a given matrix type (river water, lake
                      water, etc.)  for which the metal(s) pass the tests in Section 9.3.3, compute the
                      average percent  recovery (R) and the standard deviation of the percent
                      recovery (SR) for the metal(s). Express the accuracy assessment as a percent
                      recovery interval from R - 2SR to R + 2;SR for each matrix.  For example, if
                      R = 90% and SR = 10% for five analyses of river water, the accuracy interval
                      is expressed as 70-110%.

               9.3.4.2 Update the accuracy assessment for each metal in each matrix on a regular
                      basis (e.g., after each  five to ten new measurements).
                                                        .
9.4     Precision of matrix spike and duplicate

        9.4.1   Calculate the relative percent  difference (RPD) between the MS and MSD per the
               equation below using the concentrations found in the MS and MSD.  Do not use the
               recoveries calculated in  Section 9.3.2.4 for this calculation because the RPD is inflated .
               when the background concentration is near the spike concentration.
                                                 (Dl+D2)/2
                      Where:
                      Dl =  concentration of the analyte in  the MS sample
                      D2 =  concentration of the analyte in  the MSD sample
        9.4.2   The relative percent difference between the matrix spike and the matrix spike duplicate
               must be less than 20%. If this criterion is not met, the analytical system is be judged
             • to be out of control. In this case, correct the problem and reanalyze all samples in the
               sample batch associated with the MS/MSD that failed the RPD test.
22
January 1996

-------
                                                                                     Method 1640
 9.5     Internal standards responses—The analyst is expected to monitor the responses from the
        internal standards throughout the sample batch being analyzed. Ratios of the internal standards
        responses against each other should also be monitored routinely.  This information may be
        used to detect potential problems caused by mass dependent drift, errors incurred in adding the
        internal standards, or increases in the concentrations of individual internal standards caused by
        background contributions from the  sample.  The absolute response of any one internal standard
        must not deviate more than 60-125% of the original response in the calibration blank. If -
        deviations greater than these are observed, flush the instrument with the rinse blank and
        monitor the responses hi the  calibration blank.  If the responses of the internal standards are
        now within the limit, take a fresh aliquot of the sample, dilute by a further factor of 2, add the
        internal standards, and reanalyze. If,  after flushing, the response of the internal standards in
        the calibration blank are out  of limits, terminate the analysis and determine the cause of the
        drift.  Possible causes of drift may  be a partially blocked sampling cone or a change in the
        tuning condition of the instrument.

 9.6     Blanks—Blanks are analyzed to demonstrate freedom from contamination.

        9.6.1   Laboratory (method)  blank

               9.6.1.1 Prepare a method blank with each sample batch (samples of the same matrix
                      started through the  sample preparation process (Section 12) on the same 12-
                      hour shift, to a maximum of 10 samples). Analyze the blank immediately after
                      the OPR is analyzed (Section 9.7) to demonstrate freedom from contamination.

               9.6.1.2 If the metal of interest or any potentially interfering substance is found in the
                      blank at a concentration equal to or greater than the MDL (Table 1), sample
                      analysis must be halted, the source of the contamination determined, the
                      samples and a new  method blank prepared, and the sample batch and fresh
                      method blank reanalyzed.

               9.6.1.3 Alternatively, if a sufficient number of blanks (3 minimum) are analyzed to
                      characterize the nature of a blank, the average concentration plus two standard
                      deviations must be less than the regulatory compliance level.

               9.6.1.4 If the result for a single blank remains above the MDL or if the result for the
                      average concentration plus two standard deviations of three or more blanks
                      exceeds the regulatory compliance level, results for samples associated with
                      those blanks may not be reported  for regulatory compliance purposes.  Stated
                      another way,  results for all initial  precision and recovery tests (Section 9.2)
                      and all samples must be associated with an uncontaminated method blank
                      before these results  may be reported for regulatory compliance purposes.

        9.6.2   Field blank

               9.6.2.1 Analyze the field blank(s) shipped with each set of samples (samples collected
                      from the same site at the same time, to a maximum of 10 samples).  Analyze
                      the blank  immediately before analyzing the samples in the batch.
January 1996
23

-------
Method 1640
               9.6.2.2 If the metal of interest or any potentially interfering substance is found in the
                      field blank at a concentration equal to or Beater than the ML (Table 1) or
                      greater than one-fifth the level in the associated sample, whichever is greater,
                      results for associated samples may be the result of contamination and may not
                      be reported for regulatory compliance purposes.

               9.6.2.3 Alternatively, if a sufficient number of field blanks (3 minimum) are analyzed
                      to characterize the nature of the field blank, the average concentration plus two
                      standard deviations must be less than the regulatory compliance level or less
                      than one-half the level in the associated sample, whichever is greater.

               9.6.2.4 If contamination of the field blanks and associated samples is known or
                      suspected, the laboratory should communicate this to the sampling team so that
                      the source  of contamination can be identified and corrective measures taken
                      prior to the next sampling event.

        9.6.3   Equipment Blanks—Before any sampling equipment is used at a given site, the
               laboratory or cleaning facility is required to generate equipment blanks to demonstrate
               that the sampling equipment is free from contamination.  Two types  of equipment
               blanks are required: bottle blanks and sampler check blanks.

               9.6.3.1 Bottle blanks—After undergoing appropriate cleaning procedures (Section
                      11.4), bottles should be subjected to conditions of use to verify the
                      effectiveness of the cleaning procedures.  A representative set of sample bottles
                      should be filled with reagent water acidified to pH<2 and allowed to stand for
                      a minimum of 24 h. Ideally, the time that the bottles are allowed to stand
                      should be as close as possible to the actuid time that sample will be hi contact
                      with the bottle.  After standing, the water should be analyzed for any signs  of
                      contamination. If any bottle shows signs of contamination, the problem must
                      be identified, the cleaning procedures corrected or cleaning solutions changed,
                      and all affected bottles recleaned.

               9.6.3.2 Sampler check blanks—Sampler check blanks are generated  in the laboratory
                      or at the equipment cleaning contractor's facility by processing reagent water
                      through the sampling devices using the same procedures that are used in the
                      field (see Sampling Method). Therefore, the "clean hands/dirty hands"
                      technique used during field sampling should be followed when preparing
                      sampler check blanks at the laboratory or cleaning facility.

                      9.6.3.2.1       Sampler check blanks are generated by filling a large carboy or
                                     other container with reagent water (Section 7.2) and processing
                                     the reagent water through the equipment using the same
                                     procedures that are used iin the field  (see Sampling Method).
                                     For example, manual grab  sampler check blanks are collected
                                     by directly submerging a sample bottle into  the water, filling
                                     the bottle,  and capping. Subsurface sampler check blanks are
                                     collected by immersing the sampler into the water and
                                     pumping water into a sample container.
 24
                                                                                     January 1996

-------
                                                                                     Method 1640
                       9.6.3.2.2       The sampler check blank must be analyzed using the
                                     procedures giveffin this method.  If any metal of interest or
                                     any potentially interfering substance is detected in the blank,
                                     the source of contamination or interference must be identified
                                     and the problem corrected.  The equipment must be
                                     demonstrated to be free from the metal(s) of interest before the
                                     equipment may be used in the field.

                       9.6.3.2.3       Sampler check blanks must be run on all equipment that will
                                     be used in the field.  If, for example, samples are to be
                                     collected using both a grab sampling device and a subsurface
                                     sampling device, a sampler check blank must be run on both
                                     pieces of equipment.

9.7     Ongoing precision and recovery

        9.7.1   Prepare an ongoing precision and recovery sample (laboratory fortified method blank)
               identical to the initial precision and recovery aliquots (Section 9.2) with each sample
               batch (samples of the same matrix started through the sample preparation process
               (Section 12) on the same 12-hour shift, to a maximum of 10 samples) by spiking an
               aliquot of reagent water with the metal(s) of interest.

        9.7.2   Analyze the OPR sample before the method blank and samples from the same batch
               are analyzed.

        9.7.3   Compute the percent recovery of each metal in the OPR sample.

        9.7.4   For each metal, compare the concentration to the limits for ongoing recovery in Table
               2.  If all metals meet the acceptance criteria, system performance is acceptable and
               analysis of blanks and samples may proceed.  If, however, any individual recovery
               falls outside of the range given, the analytical processes are not being performed
               properly for that metal. Correct the problem, reprepare the  sample batch, and repeat
               the ongoing precision and recovery test (Section 9.7).

        9.7.5   Add results that pass the specifications in Section  9.7.4 to initial and previous ongoing
               data for each metal in each matrix.  Update QC charts to form a graphic representation
               of continued laboratory performance. Develop a statement of laboratory accuracy for
               each metal in each matrix type by calculating the average percent recovery (R) and the
               standard deviation of percent recovery (SR).  Express the accuracy as a recovery
               .interval from R - 2SR to R + 2SR. For example, if.R = 95% and SR =  5%, the
               accuracy is 85-105%.

9.8     The specifications contained in this method can be met if the instrument used is calibrated
        properly and then maintained in a calibrated state. A given instrument will provide the most
        reproducible results if dedicated to the settings and conditions required for the analyses of
        metals by this method.

9.9     Depending on specific program requirements, the laboratory may be required to analyze field
        duplicates  collected to determine the precision of the sampling technique. The relative percent
January 1996
25

-------
Method 1640
       difference (RPD) between field duplicates should be less than 20%.  If the RPD of the field
       duplicates exceeds 20%, the laboratory should communicate this to the sampling team so that
       the source of error can be identified and corrective measures taken before the next sampling
     .  event.

10.0  Calibration and Standardization

10.1   Operating conditions—Because of the diversity of instrument hardware, no detailed instrument
       operating conditions are provided.  The analyst is advised to follow the recommended
       operating conditions provided by the manufacturer.  The analyst is responsible for verifying
       that the instrument configuration and operating conditions satisfy the quality control
       requirements in this method.  Table 5  lists instrument operating conditions that may be used as
       a guide for analysts in determining instrument configuration and operating conditions.

10.2   Precalibration routine—The following precalibration routine should be completed before
       calibrating the instrument until it can be documented with periodic performance data that the
       instrument meets the criteria listed below without daily tuning.

    •  10.2.1  Initiate proper operating configuration of instrument and data system.  Allow a period
               of not less than 30  min for the instrument to warm up.  During this period, conduct
               mass calibration and resolution checks using the tuning solution.  Resolution at low
               mass is indicated by nickel isotopes 60, 61, 62. Resolution at high mass is indicated
               by lead isotopes 206, 207, 208.  For good performance, adjust the spectrometer
               resolution to produce a peak width of approximately 0.75 amu at 5% peak height.
               Adjust mass calibration if it has shifted by more than 0.1  amu from unit mass.

       10.2.2  Instrument stability must be demonstrated by rurnning the  tuning solution (Section 7.7)
               a minitnum of five times with resulting relative standard deviations of absolute signals
               for all analytes of less than 10%.
                                                           I
 10.3   Internal standardization—Internal standardization must be used hi all analyses to correct for
       instrument drift and physical interferences. For full mass range scans, a minimum of three
       internal standards  must be used. Internal standards must be present in all samples, standards,
       and blanks at identical levels. This may be achieved by directly adding an aliquot of the
       internal standards  to the CAL standard, blank, or sample solution (Method A), or alternatively
       by mixing with the solution before nebulization using a second channel of the peristaltic pump
       and a mixing cott (Method B). The concentration of the internal standard should be
        sufficiently high that good precision is obtained in the measurement of the isotope used for
        data  correction and to minimize the possibility of correction errors if the internal standard is
       naturally present hi the sample. Internal standards should be added to blanks, samples, and
        standards in a like way so  that dilution effects resulting from the addition may be disregarded.
        NOTE: Bismuth should not be used as an internal standard using the direct addition
        method (Method A, Section 10.3) because it is not efficiently concentrated on the
        iminodiacetate column.	

 10.4   Calibration—Before initial calibration, set up proper instrument software routines for
        quantitative analysis and connect the ICP-MS instrument to the preconcentration apparatus.
 26
January 1996

-------
                                                                                     Method 1640
        The instrument must be calibrated at a minimum of three points for each analyte to be
        determined.                      r           -

        10.4.1 Inject the calibration blank (Section 7.6.1) and calibration standards A and B (Section
               7.4.1) prepared at three or more concentrations, one of which must be at the ML
               (Table 1), and another that must be near the upper end of the linear dynamic range.
               The calibration solutions should be processed through the precpncentration system -
               using the procedures described in Section 12. A minimum of three replicate
               integrations are required for data acquisition.  Use the average of the integrations  for
               instrument calibration and data reporting.

        10.4.2 Compute the response factor at each concentration, as follows:
                                         RF =
                Where:
                Cs = concentration of the analyte  in the standard or blank solution
                Cfa = concentration of the internal standard in the solution
                As = height or area of the response at the m/z for the analyte
                A^ = height or area of the mlz for the internal standard
        10.4.3  Using the individual response factors at each concentration, compute the mean RF for
               each analyte.

        10.4.4  Linearity — If the RF over the calibration range is constant (< 20% RSD), the RF can
               be assumed to be invariant and the mean RF can be used for calculations.
               Alternatively, the results can be used to plot a calibration curve of response ratios,
                     vs. RF.
 10.5    Calibration verification — Immediately following calibration, an initial calibration verification
        should be performed.  Adjustment of the instrument is performed until verification criteria are
        met.  Only after these  criteria are met may blanks and samples be analyzed.

        10.5.1  Analyze the mid-point calibration standard (Section 10.4).

        10.5.2  Compute the percent recovery of each metal using the mean RF or calibration curve
               obtained in the initial calibration.

        10.5.3  For each metal, compare the recovery with the corresponding limit for calibration
               verification in  Table 2.  If all metals meet the acceptance criteria, system performance
               is acceptable and analysis of blanks and samples may continue using the response from
               the initial calibration. If any individual value falls outside the range given, system
               performance is unacceptable for that compound.  In this event, locate and correct the
               problem and/or prepare a new calibration check standard and repeat the test (Sections
               10.5.1-10.5.3), or recalibrate the system according to Section 10.4.
January 1996
                                                                                             27

-------
Metliod 1640
        10.5.4  Calibration must be verified following every ten samples by analyzing the mid-point
               calibration standard. If the recovery does not meet the acceptance criteria specified in
               Table 2, analysis must be halted, the problem corrected, and the instrument
               recalibrated.  All samples after the last acceptable calibration verification must be
               reanalyzed.

10.6    A calibration blank must be analyzed following every calibration verification to demonstrate
        that there is no carryover of the analytes of interest and mat the analytical system is free from
        contamination.  If the concentration of an analyte in the blank result exceeds the MDL, correct
        the problem, verify the calibration (Section 10.5), and repeat the analysis of the calibration
        blank.

11.0  Procedures for Cleaning the Apparatus

11.1    All sampling equipment, sample containers, and labware should be cleaned in a designated
        cleaning area that has been demonstrated to be free of trace element contaminants.  Such areas
        may include class 100 clean rooms as described by Moody (Reference 24), labware cleaning
        areas as described by Patterson and Settle (Reference 6), or clean benches.

11.2    Materials, such as gloves (Section 6.10.7), storage bags (Section 6.10.10), and plastic wrap
        (Section 6.10.11), may be used new without additional cleaning unless the results of the
        equipment blank pinpoint any of these materials as a source of contamination. In this case,
        either an alternate supplier must be obtained or the materials must be cleaned.

11.3    Cleaning procedures—Proper cleaning of the Apparatus is extremely important, because the
        Apparatus may not only contaminate the samples but may also remove the analytes of interest
        by adsorption onto the container surface.
        NOTE: If laboratory, field, and equipment blanks (Section 9.6) from the Apparatus
        cleaned -with fewer cleaning steps than those detailed below show no levels of analytes
        above the MDL, those cleaning steps that do not eliminate these artifacts may be
        omitted if all performance criteria outlined in Section 9 are met.	

        11.3.1   Bottles, labware, and sampling equipment

                11.3.1.1       Fill a precleaned basin (Section 6.10.8) with a sufficient quantity of a
                              0.5% solution of liquid detergent (Section 6.8), and completely
                              immerse each piece of ware.  Allow to soak in the detergent for at
                              least 30 min.

                11.3.1.2       Using a pair of clean gloves (Section 6.10.7) and clean  nonmetallic
                              brushes (Section 6.10.9), thoroughly scrub down all materials with the
                              detergent.

                11.3.1.3       Place the scrubbed materials in a precleaned basin. Change gloves.
 28
                                                                                    January 1996

-------
                                                                                      Method 1640
                11.3.1.4       Thoroughly rinse the inside and outside of each piece with reagent
                              water until there is no siffl of detergent residue (e.g., until all soap
                              bubbles disappear).

                11.3.1.5       Change gloves, immerse the rinsed equipment in a hot (50-60°C) bath
                              of concentrated reagent grade HNO3 (Section 7.1.1) and allow to soak
                              for at least 2 h.

                11.3.1.6       After soaking, use clean gloves and tongs to remove the Apparatus and
                              thoroughly rinse with distilled, deionized water (Section 7.2).

                11.3.1.7       Change gloves and immerse the Apparatus in a hot (50-60°C) bath of
                              IN trace metal grade HC1 (Section 7.1.9), and allow to soak for at
                              least 48 h.

                11.3.1.8       Thoroughly rinse all equipment and bottles with reagent water.
                              Proceed with Section 11.3.2 for labware and sampling equipment.
                              Proceed with Section 11.3.3 for sample bottles.

        11.3.2  Labware and sampling equipment

                11.3.2.1        After cleaning, air-dry in a class 100 clean air bench.

                11.3.2.2       After drying, wrap each piece of ware or equipment in two layers of
                              polyethylene film.

        11.3.3  Fluoropolymer sample bottles—These bottles should be used if mercury is a target
               analyte.

                11.3.3.1        After cleaning, fill sample bottles with 0.1% (v/v) ultrapure HC1
                              (Section 7.1.13) and cap tightly.  It may be necessary to use a strap
                              wrench to assure a tight seal.

                11.3.3.2        After capping, double-bag  each bottle in polyethylene zip-type bags.
                              Store at room temperature  until sample collection.

        11.3.4  Bottles, labware, and sampling equipment  (polyethylene  or material other than
               fluoropolymer)

               .11.3.4.1        Apply the steps outlined in Sections 11.3.1.1-11.3.1.8 to all bottles,
                              labware, and sampling equipment.  Proceed with Section 11.3.4.2 for
                              bottles or Section 11.3.4.3  for labware and sampling equipment.

               11.3.4.2        After cleaning, fill each bottle with 0.1% (v/v) ultrapure HC1 (Section
                              7.1.13).  Double-bag each  bottle in a polyethylene bag to prevent
                              contamination of the surfaces with dust and dirt.  Store at room
                              temperature until sample collection.
January 1996
                                                                                              29

-------
Method 1640
               11.3.4.3        After rinsing labware and sampling equipment, air-dry in a class 100
                             clean air bench.  After drying, wrap each piece of ware or equipment
                             in two layers of polyethylene film.
       NOTE: Polyethylene bottles cannot be used to collect samples that will be analyzed
       for mercury at trace (e.g., 0.012 jigIL) levels because of the potential for vapors tp
       diffuse through the polyethylene.	

               11.3.4.4       Polyethylene bags—If polyethylene bags need to be cleaned, clean
                             according to the following procedure:

                      11.3.4.4.1     Partially fill with cold, (1+1) HNO3 (Section 7.1.2) and rinse
                                    with distilled deionized water (Section 7.2).

                      11.3.4.4.2     Dry by hanging upside down from a plastic line with a plastic
                                    clip.

       11.3.5  Silicone tubing, fluoropolymer tubing, and other sampling apparatus—Clean any
               silicone, fluoropolymer, or other tubing used to collect samples by rinsing with 10%
               HC1 (Section 7.1.10) and flushing with water from the site before sample collection.

       11.3.6  Extension pole—Because of its length, it is impractical to submerse the 2-m
               polyethylene extension pole (used hi with the optional grab sampling device) in acid
               solutions as described above.  If such an extension pole is used, a nonmetallic  brush
               (Section 6.10.9) should be  used to scrub the pole with reagent water and the pole
               wiped down with acids described in Section 11.3.4. After cleaning, the pole should be
               wrapped in polyethylene film.

       Storage—Store each piece or assembly of the Apparatus in a clean, single polyethylene zip-
       type bag. If shipment is required,  place the bagged apparatus in a second polyethylene zip-
       type bag.

       All cleaning solutions and acid baths should be periodically monitored for accumulation of
       metals that  could lead to contamination.  When levels of metals  hi the solutions become too
       high, the solutions and baths should be changed and the old solutions neutralized and
        discarded in compliance with state and federal regulations.

 12.0  Procedures for Sample Preparation  arid Analysis
                                                           i
 12.1    Aqueous sample preparation—dissolved analytes

        12.1.1  For determination of dissolved analytes in groimd and surface waters, pipet an aliquot
               (^ 20 mL) of the filtered, acid-preserved sample into a clean 50-mL polypropylene
               centrifuge tube.  Add an appropriate volume of (1+1) nitric acid to adjust the acid
               concentration of the aliquot to approximate a 1% (v/v) nitric acid solution (e.g.,  add
               0.4 mL (1+1) HNO3 to a 20-mL aliquot of sample).  Add the internal standards, cap
               the tube, and mix. The sample is now ready for analysis. Allowance for sample
               dilution should be made in the calculations.
11.4
11.5
 30
                                                                                   January 1996

-------
                                                                                     Method 1640
 12.2   Aqueous sample preparation—total recoverable analytes
        NOTE: To preclude contamination during sample digestion, it may be necessary to
        perform the open-beaker, total-recoverable digestion procedure described in Sections
        12.2.1-12.2.6 in a fume hood that is located in a clean room. An alternate digestion
        procedure is provided in Section 12.2.7; however, this procedure has .not undergone
        interlaboratory testing.

        12.2.1  For the determination of total recoverable analytes in ambient water samples, transfer a
               100-mL (± 1 mL) aliquot from a well-mixed, acid-preserved sample to a 250-mL
               Griffin beaker (Section 6.10.3).  If appropriate, a smaller sample volume may be used.

        12.2.2  Add 2 mL (1+1) nitric acid to the beaker and place the beaker on the hot plate for
               digestion. The hot plate should be located hi a fume hood and previously adjusted to
               provide evaporation at a temperature of approximately but no higher than 85°C. (See
               the following note.)  The beaker should be covered or other necessary steps should be
               taken to prevent sample contamination from the fume hood environment.

        NOTE: For proper heating, adjust the temperature control of the hotplate so that an
        uncovered Griffin beaker containing 50 mL of water placed in the center of the hot
        plate can be maintained at a temperature approximately but no higher than 85°C.
        (Once the beaker is covered with a watch glass, the temperature of the water will rise
        to approximately 95°C.)-
        12.2.3  Reduce the volume of the sample aliquot to about 20 mL by gentle heating at 85°C.
               Do not boil.  This step takes about 2 h for a 100-mL aliquot with the rate of
               evaporation rapidly increasing as the sample volume approaches 20 mL.  (A spare
               beaker containing 20 mL of water can be used as a gauge.)

        12.2.4  Cover the lip of the beaker with a watch glass to reduce additional evaporation and
               gently reflux the sample for 30 min.  (Slight boiling may occur, but vigorous boiling
               must be avoided.)

        12.2.5  Allow the beaker to cool.  Quantitatively transfer the sample solution to a 100-mL
               volumetric flask, make to volume with reagent water, stopper, and mix.

        12.2.6  Allow any undissolved material to settle overnight, or centrifuge a portion of the
               prepared sample until clear.  (If, after centrifuging or standing overnight, the sample
               contains suspended solids that would clog the nebulizer, a portion of the  sample may
               be filtered to remove the solids before analysis.  However, care should be exercised to
               avoid potential contamination from filtration.) The sample is now ready  for analysis.
               Because the  effects of various matrices on the stability of diluted samples cannot be
            .   characterized, all analyses should be performed as soon as possible after the completed
               preparation.

        12.2.7  Alternate total recoverable digestion procedure
January 1996
                                                                                             31

-------
Method 1640
               12.2.7.1       Open the preserved sample under clean conditions. Add ultrapure
                             nitric acid at the rate of 10 mL/L.  Remove the cap from the original
                             container only long enough to add the aliquot of acid. The sample
                             container should not be filled to the lip by the addition of the acid.
                             However, only minimal headspace is needed to avoid leakage during
                             heating.

               12.2.7.2       Tightly recap the container and sihake thoroughly.  Place the container
                             in an oven preheated to 85°C. The container should be placed on an
                             insulating piece of material such as wood rather than directly on the
                             typical metal grating. After the samples have reached 85°C, heat for  2
                             h. (Total time will be 2.5-3 h depending on the sample size).
                             Temperature can be monitored using an identical sample container with
                             distilled water and a thermocouple to standardize heating tune.

               12.2.7.3       Allow the sample to cool.  Add the internal standards and mix.  The   ;
                             sample is now ready for analysis.  Remove aliquots for analysis under
                             clean conditions.

12.3    Before first use, the preconcentration system should be thoroughly cleaned and decontaminated
        using 0.2M oxalic acid.

        12.3.1  Place approximately 500 mL 0.2M oxalic acid in all the eluent/solution containers and
               fill the sample loop with 0.2M oxalic acid using the sample pump (P4) at a flow rate
               of 3-5 rnL/min.  With the preconcentration system disconnected from the ICP-MS
               instrument, use the pump program sequence listed in Table 3, to flush the complete
               system with oxalic acid. Repeat the flush sequence three times.

        12.3.2  Repeat the sequence described in Section 12.3.1 using  1.25M nitric acid and again
               using reagent water hi place of the 0.2M oxalic acid.

        12.3.3  Rinse the containers thoroughly with reagent water, fill them with their designated
               reagents (see Figure 1), and run through the sequence in Table 3 once to prime the
               pump and all eluent lines with the correct reagents.

 12.4    Sample Analysis
                                                          j
        12.4.1  Initiate ICP-MS instrument operating configuration.  Tune the instrument for the
               analytes of interest (Section 10).

        12.4.2 Establish instrument software run procedures foir quantitative analysis. Because the
               analytes are eluted from the preconcentration column in a transient manner, it is
               recommended that the instrument software be configured hi a rapid scan/peak hopping
               mode. The instrument is now ready to be calibrated.

        12.4.3- Reconnect the preconcentration system to the IC3P-MS instrument. With valves A and
               B hi the off position and valve C hi the on position, load the sample through the
               sample loop to waste using pump P4 for 4 min at 4 mL/min.  Switch on the carrier
 32
January 1996

-------
                                                                                   Method 1640
               pump (P3) and pump 1% nitric acid to the nebulizer of the ICP-MS instrument at a
               flow rate of 0.8-1.0 mL/miri."

        12.4.4  Switch on the buffer pump (P2), and pump 2M ammonium acetate at a flow rate of
               1.0 mL/min.

        12.4.5  Preconcentration of the sample may be achieved by running through an eluent pump
               program (PI) sequence similar to that illustrated in Table 3.  The exact timing of this
               sequence should be modified according to the internal volume of the connecting tubing
               and the specific hardware configuration used.

               12.4.5.1        Inject sample—With valves A, B and C on, load sample from the loop
                             onto the column using 1M ammonium acetate for 4.5 min at 4.0
                             mL/min. The analytes are retained on the column, while most of the
                             matrix is passed through to waste.

               12.4.5.2        Elute analytes—Turn off valves A and B  and begin eluting the
                             analytes by pumping 1.25M nitric acid through the column at 4.0
                             mL/min, then turn off valve C and pump  the  eluted analytes into the
                             ICP-MS instrument at 1.0  mL/min.  Initiate ICP-MS software data
                             acquisition and integrate the eluted analyte profiles.

               12.4.5.3        Column Reconditioning—Turn on valve C to direct column effluent to
                             waste, and pump 1.25M nitric acid,  1M ammonium acetate, 1.25M
                             nitric acid and  1M ammonium acetate alternately through the column
                             at 4.0 mL/min. During this process, the next sample can be loaded
                             into the sample loop using the sample pump (P4).

        12.4.6  Repeat the sequence described  in Section 12.4.5 for each  sample to be analyzed. At
               the end of the analytical run, leave the column filled with 1M ammonium acetate
               buffer until it is next used.

        12.4.7  Samples  having concentrations  higher than the established linear dynamic range should
               be diluted into range  with 1% HNO3 (v/v) and reanalyzed.


13.0   Data Analysis and Calculations

13.1   Elemental equations recommended for sample data calculations are listed in Table 4.  Sample
       data should be reported in units of ug/L (parts-per-billion; ppb).  Report results at or above the
       ML for metals found in samples and determined in standards.  Report all results for metals
       found in blanks,  regardless of level.

13.2   For data values less than the ML, two significant figures should be used for reporting element
       concentrations.  For data values greater than or equal  to the ML, three significant figures
       should be used.

13.3   Compute the concentration of each analyte in the sample using the response factor determined
       from calibration data (Section 10.4) and the following equation:
January 1996
33

-------
Method 1640
                                     Cs (mgIL) =
                        Where  the terms are as defined in Section 10.4.2.
13.4   Corrections for characterized spectral interferences should be applied to the data.  Chloride
       interference corrections should be made on all samples, regardless of the addition of
       hydrochloric acid, because the chloride ion is a common constituent of environmental samples.

13.5   If an element has more than one monitored m/z, examina.tion of the concentration calculated
       for each m/z, or the relative abundances, will provide useful information for the analyst in
       detecting a possible spectral interference.  Consideration should therefore be given to both
       primary and  secondary m/z's in the evaluation of the element concentration. In some cases,
       the secondary m/z may be less sensitive or more prone to interferences than the primary
     .  recommended m/z;  therefore, differences between the results do not necessarily indicate a
       problem with data calculated for the primary m/z.

13.6   The QC data obtained during the analyses provide an indication of the quality of the sample
       data and should be  provided with the sample results.

13.7   Do not perform blank subtraction on the sample results. Report results for samples and
       accompanying blanks.

14.0 Method Performance

14.1   The method  detection limits (MDLs) listed in Table 1 and the quality control acceptance
       criteria listed in Table 2 were validated in two laboratories (Reference 25) for dissolved
       analytes.


15.0  Pollution Prevention

 15.1   Pollution prevention encompasses any technique that reduces or eliminates the quantity or
        toxicity of waste at the point of generation.  Many oppoirtunities for pollution prevention exist
        in laboratory operation. The EPA has  established a preferred hierarchy of environmental
        management techniques that places pollution prevention as the management option of first
        choice.  Whenever feasible, laboratory personnel should use pollution prevention techniques to
        address their waste generation. When  wastes cannot be feasibly reduced at the source, the
        Agency recommends recycling as the next best option.  The acids used in this method should
        be reused as practicable by purifying by electrochemical techniques.  The only other chemicals
        used in this  method are the neat materials used in preparing standards.  These standards are
        used in extremely small amounts and pose little threat to the environment when managed
        properly.  Standards should be prepared in volumes consistent  with laboratory use to minimize
        the volume of expired standards to be  disposed.
 34
January 1996

-------
                                                                                 Method 1640
 15.2    For information about pollution prevention that may be applicable to laboratories and research
        institutions, consult Less is Better: Laboratory Chemical Management for Waste Reduction,
        available from the American Chemical Society's Department of Government Relations and
        Science Policy, 1155 16th Street NW, Washington DC 20036, 202/872-4477.

 16.0  Waste  Management

 16.1    The Environmental Protection Agency requires that laboratory waste management practices be
        conducted consistent with all applicable rules and regulations. The Agency urges laboratories
        to protect the air, water, and land by minimizing and controlling all releases from hoods and
        bench operations, complying with the letter and spirit of any sewer discharge permits and
        regulations, and by complying with all solid and hazardous waste regulations, particularly the
        hazardous waste identification rules and land disposal restrictions.  For further information on
        waste management consult The Waste Management Manual for Laboratory Personnel,
        available from the American Chemical Society at the address listed in the Section 15.2.
17.0  References

1       Adeloju, S.B.; Bond, A.M. "Influence of Laboratory Environment on the Precision and
        Accuracy of Trace Element Analysis," Anal. Chem. 1985, 57, 1728.

2       Berman, S.S.; Yeats, P.A. "Sampling of Seawater for Trace Metals," CRC Reviews in
       Analytical Chemistry 1985, 16, 1.

3       Bloom, N.S. "Ultra-Clean Sampling, Storage, and Analytical  Strategies for the Accurate
        Determination of Trace Metals in Natural Waters"; Presented at the 16th Annual EPA
        Conference on the Analysis of Pollutants in the Environment, Norfolk, VA, May 5, 1993.

4       Bruland, K.W. "Trace Elements hi Seawater," Chemical Oceanography 1983, 8, 157.

5       Nriagu, J.O.; Larson, G.; Wong, H.K.T.; Azcue, J.M. "A Protocol for Minimizing
        Contamination in the Analysis of Trace Metals hi Great Lakes Waters," /. Great Lakes
       Research 1993, 19, 175.

6      Patterson, C.C.; Settle, D.M. "Accuracy in Trace Analysis"; In National Bureau of Standards
        Special Publication 422; LaFleur, P.D., Ed., U.S. Government Printing Office, Washington,
        DC, 1976.

7      Fitzgerald, W.F.; Watras, CJ. Science of the Total Environment 1989, 87/88, 223.

8       Gill, G.A.; Fitzgerald, W.F. Deep Sea Res. 1985, 32,  287.

9      Prothro, Martha G., "Office of Water Policy and Technical Guidance on Interpretation and
       Implementation of Aquatic Life Metals Criteria," EPA Memorandum to Regional Water
        Management and Environmental Services Division Directors, Oct. 1, 1993.
January 1996
35

-------
Method 1640
10     "Format for Method Documentation", Distributed by the EPA Environmental Monitoring
       Management Council, Washington, DC, Nov. 18, 1993.

11   .  Siraraks, A.; Kingston, H.M.; Riviello, J.M. Anal Chem. 1990, 62, 1185.

12     Heithmar, E.M., Hinners, T.A.; Rowan, J.T.; Riviello, J.M. Anal Chem. 1990, 62, 857.

13     Gray, A.L.; Date, A.R. Analyst, 1983,108, 1033.

14     Houk, R.S. et al. Anal Chem. 1980, 52, 2283.          i
                                                         i
15     Houk, R.S. Anal.  Chem.  1986, 58, 97A.

16     Windom, H.L; Byrd, J.T.; Smith, R.G., Jr.; Huan, F. "Inadequacy of NASQAN Data for
       Assessing Metal Trends in the Nation's Rivers," Environ.  Sci. Technol. 1991, 25, 1137.

17     Zief, M.; Mitchell, J.W.  "Contamination Control in Trace  Metals Analysis"; In Chemical
       Analysis 1976, Vol. 47, Chapter  6.

18     Thompson, JJ.; Houk, R.S. Appl. Spec. 1987, 41, 801.  :                                   \

19     Carcinogens - Working With Carcinogens; Department of Health, Education, and Welfare,
       Public Health Service. Centers for Disease Control. National Institute for Occupational Safety
       and Health Publication No. 77-206, Aug.  1977.  Available from the National Technical
       Information Service (NTIS) as PB-277256.

20     "OSHA Safety and Health Standards, General Industry," 29 CFR 1910, Occupational Safety
       and Health Administration, OSHA 2206 (revised January  1976).

21     Safety in Academic Chemistry Laboratories, American  Chemical Society Publication,         ,
       Committee on Chemical Safety,  3rd Edition, 1979.

22     "Proposed OSHA Safety and Health Standards, Laboratories, Occupational Safety and Health
       Administration," Fed. Regist., July 24, 1986.

23     Handbook of Analytical  Quality  Control in Water and Wastewater Laboratories; U.S.
       Environmental Protection Agency. EMSL-Cincinnati, OH, March 1979; EPA-600/4-79-019.

24     Moody, J.R. "NBS Clean Laboratories for Trace Element Analysis," Anal. Chem. 1982, 54,
       1358A.
                                                         :                               "    i
25     "Results of the Validation Study for Determination of Trace Metals at EPA Water Quality
       Criteria Levels," April 1995. Available from the Sample  Control Center (operated by
       DynCorp), 300 N. Lee Street, Alexandria, VA 22314, 703/519-1140.
 36
January 1996

-------
                                                                                     Method 1640
 18.0  Glossary

        Many of the terms and definitions listed below are used in the EPA 1600-series methods, but
        terms have been cross-referenced to terms commonly used in other methods where possible.

 18.1   Ambient Water—Waters in the natural environment (e.g., rivers, lakes, streams, and other"
        receiving waters), as opposed to effluent discharges.

 18.2   Analyte—A metal tested for by the methods referenced in this method. The analytes are
        listed in Table 1.
 18.3
 18.4
 18.5
 18.6
 18.7
 18.8
18.9
 Apparatus—The sample container and other containers, filters, filter holders, labware, tubing,
 pipets, and other materials and devices used for sample collection or sample preparation, and
 that will contact samples, blanks, or analytical standards.

 Calibration Blank—A volume of reagent water acidified with the same acid matrix as in the
 calibration standards. The calibration blank is a zero standard and is used to calibrate the ICP
 instrument (Section 7.6.1).

 Calibration Standard (CAL)—A solution prepared from a dilute mixed standard and/or stock
 solutions and used to calibrate the response of the instrument with respect to analyte
 concentration.

 Dissolved Analyte—The concentration of analyte in an aqueous sample that will pass through
 a 0.45-um membrane filter assembly prior to sample acidification (Section 8.3).

 Equipment Blank—An aliquot of reagent water that is subjected in the laboratory to all
 aspects of sample collection and analysis, including contact with all sampling devices and
 apparatus.  The purpose of the equipment blank is to  determine if the sampling devices and
 apparatus for sample collection have been adequately cleaned before they are shipped to the
 field site. An acceptable equipment blank must be achieved before the sampling devices and
 apparatus are used for sample collection.  In addition, equipment blanks should be run on
 random, representative sets of gloves, storage bags, and plastic wrap for each lot to determine
 if these materials are free from contamination before they are used.

 Field Blank—An aliquot of reagent water that is placed in a sample container in the
 laboratory, shipped to the field, and treated  as a sample in all respects, including contact with
 the sampling devices and exposure to sampling site conditions, storage, preservation, and all
 analytical procedures, which may include filtration. The purpose  of the field blank is to
 determine if the field or sample transporting procedures and environments have contaminated
 the sample.

 Field Duplicates (FD1 and FD2)—Two separate samples collected in separate sample bottles
 at the same time and place under identical circumstances and treated exactly the same
 throughout field and laboratory procedures.  Analyses of FD1 and ED2 give a measure of the
precision associated with sample collection, preservation, and storage, as well as with
laboratory procedures.
January 1996
                                                                                             37

-------
Method 1640
18.10  Initial Precision and Recovery (JPR)—Four aliquots of the OPR standard analyzed to
       establish the ability to generate acceptable precision and accuracy. IPRs are performed before
       the first time a method is used and any time the method or instrumentation is modified.

18.11  Instrument Detection Limit (IDL)—The concentration equivalent to the analyte signal which
       is equal to three times the standard deviation of a series of ten replicate measurements of the
       calibration blank signal at the selected analytical masses).

18.12  Internal Standard—Pure analyte(s) added to a sample, extract, or standard solution in known
       amount(s) and used to measure the relative responses of other method analytes that are
       components of the same sample or solution. The internal standard must be an analyte that is
       not a sample component (Sections 7.5 and 9.5).

18.13  Laboratory Blank—An aliquot of reagent water that is treated exactly as a sample including
       exposure to all glassware, equipment, solvents,  reagents, internal standards, and surrogates that
       are used with samples.   The laboratory blank is used to determine if method analytes or
       interferences are present in the laboratory environment, the reagents, or the apparatus (Sections
       7.6.2 and 9.6.1).

18.14  Laboratory Control Sample (LCS>—See Ongoing Precision and Recovery (OPR) Standard.

18.15  Laboratory Duplicates (LD1 and LD2)—Two aliquots; of the same sample taken hi the
       laboratory and analyzed separately with identical procedures. Analyses of LD1 and LD2
       indicates precision associated with laboratory procedures, but not with sample collection,
       preservation, or storage procedures.
                                                         I
18.16  Laboratory Fortified Blank (LFB>—See Ongoing Precision and Recovery (OPR) Standard.

18.17  Laboratory Fortified Sample Matrix (LFM)—See Matrix Spike (MS) and Matrix Spike
       Duplicate (MSD).

18.18  Laboratory Reagent Blank (LRB)—See Laboratory Blank.

 18.19  Linear Dynamic Range (LDR)—The concentration range over which the instrument response
       to an analyte is linear (Section 9.2.3).                                                   ;

 18.20  Matrix Spike (MS) and Matrix Spike Duplicate (MSD)—Aliquots of an environmental
        sample to which known quantities of the method analytes are added hi the laboratory. The
        MS and MSD are analyzed exactly like a sample. Their purpose is to quantify the bias and
        precision caused by the sample matrix.  The background concentrations of the analytes hi the
        sample matrix must be determined hi a separate aliquot and the measured values hi the MS
        and MSD corrected for background concentrations (Section 9.3).

 18.21   m/z—mass-to-charge ratio                          i

 18.22  May—This action, activity, or procedural step is optional.

 18.23  May Not—This action, activity, or procedural step is prohibited.
                                                                                  January 1996

-------
                                                                                    Method 1640
 18.24  Method Blank—See Laboratory Blank.

 18.25  Method Detection Limit (MDL)—The minimum concentration of an analyte that can be
        identified, measured, and reported with 99% confidence that the analyte concentration is
        greater than zero (Section 9.2.1 and Table 1).

 18.26  Minimum Level (ML)—The lowest level at which the entire analytical system gives a
        recognizable signal and acceptable calibration point (Reference 9).

 18.27  Must—This action, activity, or procedural step is required.

 18.28  Ongoing Precision and Recovery (OPR) Standard—A laboratory blank spiked with known
        quantities of the method analytes.  The OPR is analyzed exactly like a sample.  Its purpose is
        to determine whether the methodology is hi control and to assure that the results produced by
        the laboratory remain within the method-specified limits for precision and accuracy (Sections
        7.9 and 9.7).

 18.29  Preparation Blank—See Laboratory Blank.

 18.30  Primary Dilution Standard—A solution containing the analytes that is purchased or prepared
        from stock solutions and diluted as needed to prepare calibration solutions and other solutions.

 18.31   Quality Control Sample (QCS)—A sample containing all or a subset of the method analytes
        at known concentrations.  The QCS is obtained from a source external to the laboratory or is
        prepared from a source of standards different from the source of calibration standards. It is
        used to check laboratory performance with test materials prepared external to the normal
        preparation process.

 18.32  Reagent Water—Water demonstrated to be free from the method analytes and potentially
        interfering substances at the MDL for that metal in the method.

 18.33   Should—This action, activity, or procedural  step is suggested but not required.

 18.34  Stock Standard Solution—A solution containing one or more method analytes that is
        prepared using a reference material traceable to EPA, the National Institute of Science and
        Technology (NIST), or a source that will attest to the purity and authenticity of the reference
        material.

 18.35   Total Recoverable Analyte—The concentration of analyte determined by  analysis of the
        solution extract of an unfiltered aqueous sample following digestion by refluxing with hot
        dilute mineral acid(s) as specified in the method (Section 12.2).

 18.36   Tuning Solution—A solution used to determine acceptable instrument performance before
        calibration and sample analyses (Section 7.7).
January 1996
39

-------
Method 1640
                                                         Table 1
                                                                             i
   List of Analytes Amenable to Analysis Using Method 1640:  Lowest Water Quality Criterion
                   for Each Metal Species, Method Detection Limits, Minimum Levels,
                                      and Recommended Analytical Masses


Metal

Cadmium
Copper
Lead
Nickel
Lowest
Ambient
Water Quality
Criterion
(ug/L)1
0.37
2.5
0.54
8.2
Method Detection Limit
(MDL) and Minimum
Level (ML); ugflL
MDL2
0.0024
0.024
0.0081
0.029
ML3
0.01
0.1
0.02
0.1

Recommended
Analytical m/z

111
63
206, 207, 208
60
 2.

 3.
Lowest of the freshwater, marine, or human health WQC at 40 CFR Part 131 (57 FR 60848 for human health criteria and 60 FR 22228 for aquatic criteria).
Hardness^lependent freshwater aquatic life criteria also calculated to reflect a hardness ol 25 mg/L CaCOj, and all aquatic life criteria adjusted to reflect
dissolved levels in accordance with equations provided ia60FR 22228. Hardnesiwlependent dissolved criteria conversion factors for Cd and Pb also calculated
at a hardness of 25 mg/L per 60 FR 22228.

Method Detection Limit as determined by 40 CFR Part 136, Appendix B.

Minimum Level (ML) calculated by multiplying laboratory-determined MDL by 3.18 and rounding result to nearest multiple of 1,2,5,10, etc. in accordance
with procedures used by BAD and described in the EPA Draft National Guidance for the Permitting, Monitoring, and Enforcement of Water Quality-Based
Effluent limitations Set Below Analytical DetecaonfQiamtitation Levels, March 22,1994.
 40
                                                                                                    January 1996

-------
                                                                   Method 1640
 TABLE 2: QUALITY CONTROL ACCEPTANCE CRITERIA FOR PERFORMANCE TESTS
                            IN EPA METHOD 16401
Metal
Cadmium
Copper
Lead
Nickel
Initial Precision and
Recovery (Section 9.2)
s X
23 75-121
43 67-154
44 56-144
27 74-128
Calibration Verification
(Section 10.5)
86-110
77-119
78-122
87-115
Ongoing Precision and
Recovery (Section 9.7)
73-123
63-159
52-144
71-130
Spike Recovery
(Section 93)
73-123
63-159
52-144
71-130
1.
All specifications expressed as percent
January 1996
                                                                    41

-------
Method 1640
  TABLE 3: ELTJENT PUMP PROGRAMMING SEQUENCE FOR PRECONCENTRATION
                               OF TRACE ELEMENTS
                     Time     Flow
                     (min)   (mL/min)
                                          Eluent
Valve   Valve
 A,B     C •
                     0.0      4.0     1M ammonium acetate    ON    ON

                     4.5      4.0        1.25MHNO3       ON    ON
                                                    I
                     5.1      1.0        1.25MHNO3       OFF    ON

                     5.5      1.0        1.25MHNO3       OFF    OFF

                     7.5      4.0        1.25MHNO3       OFF    ON

                     8.0      4.0     1M ammonium acetate    OFF    ON

                     10.0     4.0        1.25MHNO3       OFF    ON

                     11.0     4.0     1M ammonium acetate    OFF    ON

                     125     0.0                        OFF    ON
 42
                     January 1996

-------
                                                                           Method 1640
  TABLE 4: RECOMMENDED ANALYTICAL ISOTOPES AND ELEMENTAL EQUATIONS
                                      FOR DATA CALCULATIONS
Element
Cd
Cu
Pb
Ni
Isotope
106.108.111.114
63,65
206.207.208
60
Elemental Equation . Note
(1.000)(111C)-(1.073)[(108C)-(0.712)(106C)] (1)
(1.000)(63C)
(1.000)(20SC)+(1.000)(207C)+(1.000)(208C) (2)
(l.OOOX^C)
  C—counts at specified m/Z.
  (1)—correction for MoO interference. An additional isobaric elemental correction should be made if
  palladium is present.
  (2)—allowance for variability of lead isotopes.

  NOTE: As a minimum, all isotopes listed should be monitored. Isotopes recommended for analytical
  determination are underlined.
January 1996
43

-------
Method 1640
        TABLE 5: RECOMMENDED INSTRUMENTAL OPERATING CONDITIONS
         Chromatography

         Instrument
         Preconcentration column
Dionex chelation system
Dionex MetPac CC-1
        ICP-MS Instrument Conditions

        Instrument
        Plasma forward power
        Coolant flow rate
        Auxiliary flow rate
        Nebulizer flow rate

        Internal standards

        Data Acquisition

        Detector mode
        Mass range
        Dwell time
        Number of MCA channels
        Number of scan sweeps
VG PlasmaQuad Type I
1.35 kW
13.5L/min
0.6 L/min
0.78 L/min

Sc, Y, In, Tb
Pulse counting
45-240 amu
160 ps
2048
250
44
                                 January 1996

-------
                                                                            Method 1640
                   Figure 1. Configuration of Preconcentration System
                                      1MNH4OAc 1.25 M Nitric Acid
 2M NH4OAc
                                                               Mixing Tee
                                                                               62-001-63
January 1996
                                                                                   45

-------

-------