United States
        Environmental Protection
        Agency	
Office of Water
Washington, DC 20460
EPA-821-R-98-002
  PB99-121949
  February 1999
SEPA  Method 1664, Revision A:
        N-Hexane Extractable Material (HEM; Oil
        and Grease) and Silica Gel Treated
        N-Hexane Extractable Material (SGT-
        HEM; Non-polar Material) by Extraction
        and Gravimetry
                                            ~cra.MK.cWw

-------
                                Acknowledgments

   This method was prepared under the direction of William A. Telliard of the Engineering and
Analysis Division within EPA's Office of Water. This document was prepared under EPA Contract
                                   No. 68-C-98-139
                    by DynCorp Information and Enterprise Technology.
                                     Disclaimer

  This method has been reviewed by the Engineering and Analysis Division, U.S. Environmental
Protection Agency, and approved for publication. Mention of trade names or commercial products
               does not constitute and endorsement or recommendation for use.

-------
Table of Contents




Introduction	  iv




1.0 Scope and Application	1




2.0 Summary of Method 	2




3.0 Definitions  	2




4.0 Interferences  	2




5.0 Safety	2




6.0 Equipment and Supplies  	3




7.0 Reagents and Standards  	5




8.0 Sample Collection, Preservation, and Storage	6




9.0 Quality Control	7




10.0           Calibration and Standardization	14




11.0           Procedure  	14




12.0           Data Analysis and Calculations	19




13.0           Method Performance	20




14.0           Pollution Prevention 	20




15.0           Waste Management	20




16.0           References	20




17.0           Tables	22




18.0           Glossary of Definitions and Purposes	22

-------
Method 1664
Introduction

As a party to the Montreal Protocol on Substances that Deplete the Ozone Layer and as required by
law under the Clean Air Act Amendments of 1990 (CAAA), the United States is committed to
controlling and eventually phasing out the use of chlorofluorocarbons (CFCs). In support of these
efforts, Method 1664 was developed by the United States Environmental Protection Agency Office of
Science and Technology to replace previously used gravimetric procedures that employed Freon-113,
a Class I CFC, as the extraction solvent for the determination of oil and grease and non-polar
material.

Method 1664 is a performance-based method applicable to aqueous matrices that requires the use of
n-hexane as the extraction solvent and gravimetry as the determinative technique. Alternative
extraction and concentration techniques are allowed, provided that all performance specifications are
met. In addition, QC procedures designed to monitor precision and accuracy have been incorporated
into Method 1664.

Questions concerning this method should be directed to:

William A. Telliard
U.S. EPA
Engineering and Analysis Division, MC 4303
Office of Science and Technology
401 M Street, SW
Washington, D.C.  20460

Requests for additional copies of this method should be direct to:

National Technical Information Service
5285 Port Royal Road
Springfield, VA 22161
Phone: (703)487-4650
       (800) 553-6847
Document No. PB99-121949

U.S. EPA National Center for Environmental
       Publications and Information
11029 Kenwood Road
Cincinnati, OH 45242
Phone: (513)489-8190
       (800) 490-9198
Document No. EPA-821-R-98-002
iv                                                                               February 1999

-------
                                                                                      Method 1664
February 1999

-------
                                                                            Method 1664
                            Method 1664, Revision A
   n-Hexane Extractable Material (HEM; Oil and Grease) and Silica Gel
 Treated n-Hexane Extractable Material (SGT-HEM;  Non-polar Material)
                         by Extraction and Gravimetry


1.0    Scope and Application
 1.1    This method is for determination of n-hexane extractable material (HEM; oil and grease) and
       n-hexane extractable material that is not adsorbed by silica  gel (SGT-HEM; non-polar
       material) in  surface and saline  waters  and  industrial and  domestic  aqueous wastes.
       Extractable  materials that may be determined are relatively  non-volatile hydrocarbons,
       vegetable oils, animal fats, waxes, soaps, greases, and related materials. The method is based
       on prior Environmental Protection Agency (EPA) methods for determination of "oil and
       grease" and  "total petroleum hydrocarbons" (References 16.1 and 16.2).
 1.2    This method is for use in the Environmental  Protection  Agency's  (EPA's)  survey and
       monitoring programs under the Clean Water Act; the Resource  Conservation and Recovery
       Act; the Comprehensive Environmental Response, Compensation, and Liability Act; and other
       EPA regulatory programs. "Oil and grease" is a conventional pollutant under the Clean Water
       Act and codified at 40 CFR 401.16.  The term "n-hexane extractable material" reflects that
       this method can used to determine materials other than oils and  greases. Similarly, the term
       "silica gel treated n-hexane extractable  material" reflects that  this method can be used to
       determine material that is not adsorbed by silica gel (non-polar  material).
 1.3    This method is not applicable to measurement of materials that volatilize at temperatures
       below approximately 85 °C.  Petroleum fuels from gasoline through #2 fuel oil may be
       partially lost in the solvent removal operation.
 1.4    Some crude  oils and heavy fuel oils contain a significant percentage of materials that are not
       soluble in n-hexane. Accordingly, recoveries of these materials may be low.
 1.5    This method is capable of measuring HEM and SGT-HEM in the range of 5 to 1000 mg/L,
       and may be extended to higher levels by analysis of a smaller sample  volume collected
       separately.
 1.6    For HEM and SGT-HEM in this method, the method detection limit (MDL) is 1.4 mg/L and
       the minimum level of quantitation (ML) is 5.0 mg/L (Reference 16.3).
 1.7    The laboratory is permitted to modify the method to overcome interferences or lower the cost
       of measurements, provided that all performance criteria in this method are met.  The
       requirements for establishing method equivalency are given in  Sections 9.1.2 and 9.2.3.
 1.8    Any modification of this method, beyond those expressly permitted, shall be considered a
       major modification subject to application and approval of alternate test procedures under 40
       CFR 136.4 and 136.5.
 1.9    Each laboratory that uses this method must demonstrate the ability to generate acceptable
       results using the procedure in Section 9.2.
February 1999

-------
Method 1664
2.0    Summary of Method
 2.1    A 1-L sample is acidified to pH <2 and serially extracted three times with n-hexane in a
       separately funnel. The extract is dried over sodium sulfate.
 2.2    The solvent is distilled from the extract and the HEM is desiccated and weighed. If the HEM
       is to be used for determination of SGT-HEM, the HEM is redissolved in n-hexane.
 2.3    For SGT-HEM determination, an amount of silica gel proportionate to the amount of HEM
       is added to the solution containing the redissolved HEM to remove polar materials. The
       solution is filtered to remove the silica  gel, the solvent is distilled, and the SGT-HEM is
       desiccated and weighed.
 2.4    Quality  is  assured  through  calibration and testing of  the extraction, distillation, and
       gravimetric systems.

3.0    Definitions
 3.1    HEM and SGT-HEM are method-defined analytes; i.e., the definitions of both HEM and
       SGT-HEM are dependent on the procedure used. The nature of the oils and/or greases, and
       the presence of extractable non-oily matter in the sample will influence the material measured
       and interpretation of results.
 3.2    Definitions for terms used in this method are given in the glossary at the end of the method.

4.0    Interferences
 4.1    Solvents, reagents, glassware, and other sample-processing hardware may yield artifacts that
       affect results.  Specific selection of reagents and purification of solvents may be required.
 4.2    All materials  used in the analysis  shall  be demonstrated  to be free from interferences  by
       running laboratory blanks as described in Section 9.4.
 4.3    Glassware is  cleaned by washing in hot water containing detergent, rinsing  with tap and
       distilled water, and  rinsing with solvent or baking.  Boiling flasks that will contain the
       extracted residue are dried in an oven at 105-115°C and stored in a desiccator.
 4.4    Sodium sulfate and silica gel fines have the potential to inflate results for HEM and SGT-
       HEM by passing through the filter paper. If the filter paper specified in this method is
       inadequate for removal of these fines, use of a 0.45-micron filter is recommended.
 4.5    Interferences extracted from samples will vary considerably from source to source, depending
       upon the diversity of the site being sampled. For those instances in which samples are thought
       to consist of complex matrices containing substances (such as particulates or detergents) that
       may interfere with the extraction procedure, a smaller sample may need to be collected for
       analysis.

5.0    Safety
 5.1    The toxicity or carcinogen!city of each reagent used in  this method has not been precisely
       determined; however, each chemical should be treated as a potential health hazard. Exposure
       to these chemicals should be reduced to the lowest possible level. It is suggested that the
       laboratory perform personal hygiene monitoring of each analyst that uses this method. This
       monitoring should be performed using Occupational  Safety and Health Administration
       (OSHA) or National Institute of Occupational Safety and Health (NIOSH) approved personal

2                                                                             February 1999

-------
                                                                         Method 1664
      hygiene monitoring methods. Results of this monitoring should be made available to the
      analyst.
 5.2   n-Hexane has been shown to have increased neurotoxic effects over other hexanes and some
      other solvents. OSHA has proposed a time-weighted average (TWA) of 50 parts-per-million
      (ppm); NIOSH concurs that an 8-hour TWA/permissible exposure limit (PEL) of 50 ppm is
      appropriate for  n-hexane;  and the American  Conference of Governmental Industrial
      Hygienists (ACGIH) has published a threshold limit value (TLV) of 50 ppm for n-hexane.
      Inhalation of n-hexane should be minimized by performing all operations with n-hexane in an
      explosion-proof hood or well-ventilated area.
 5.3   n-Hexane has a flash point of-23 °C (-9 °F), has explosive limits in air in the range of 1 - 7
      percent, and poses a serious fire risk when heated or exposed to flame. n-Hexane can react
      vigorously with oxidizing  materials.  The laboratory  should  include  procedures in its
      operations that address the safe handling of n-hexane.
 5.4   Unknown samples may contain high concentrations of volatile toxic compounds.  Sample
      containers should be opened in a hood and handled with gloves to prevent exposure.
 5.5   This method does not address all  safety issues associated with  its use.  The laboratory is
      responsible for maintaining a safe work environment and a current awareness file of OSHA
      regulations regarding the safe handling of the chemicals specified in this method. A reference
      file of material safety data sheets (MSDSs) should be available to all personnel involved in
      these analyses.  Additional information on laboratory safety can be found in References
      16.4-16.6.

6.0   Equipment and  Supplies

   NOTE:  Brand names,  suppliers, and part  numbers  are  for illustrative
   purposes only.   No  endorsement  is implied.  Equivalent performance
   may  be achieved using   apparatus  and  materials other  than those
   specified here,   but  demonstration  of equivalent performance that
   meets the requirements  of  this method  is  the responsibility of  the
   laboratory.

 6.1   Sampling equipment.
     6.1.1    Sample collection bottles—Glass, approximately 1-L, with PTFE-lined screw cap.

   NOTE: In those instances necessitating  collection of a sample known
   or suspected to  contain  >500 mg/L of HEM  (Section 8.1.2),  a smaller
   sample container may be  used.
     6.1.2    Cleaning.
          6.1.2.1   Bottles—Detergent water wash, tap water rinse, cap with aluminum foil, and
                   bake at 200-250 °C for 1 h minimum prior to use. Solvent rinse may be used
                   in place of baking.
          6.1.2.2   Liners for screw caps—Detergent water wash, tap water and solvent rinse,
                   and bake at 110-200 °C for 1 h minimum prior to use.
February 1999

-------
Method 1664
      6.1.3    Bottles and liners must be lot-certified to be free of artifacts by running laboratory
              blanks according to this method (per Section 9.4). If blanks from bottles and/or liners
              without cleaning or with fewer cleaning steps than required above show no detectable
              materials, the bottle and liner cleaning steps that do not eliminate these artifacts may
              be omitted.
6.2    Equipment for glassware cleaning.
      6.2.1    Laboratory sink with overhead fume hood.
      6.2.2    Oven—Capable of maintaining a temperature within ± 2° C in the range of 70-250 °C.
6.3    Equipment for calibration.
      6.3.1    Analytical Balance—Capable of weighing 0.1 mg.
      6.3.2    Volumetric flask—Glass, 100-mL.
      6.3.3    Vials—Assorted sizes, with PTFE-lined screw caps.
      6.3.4    Volumetric pipette—Glass, 5-mL.
6.2    Equipment for sample extraction.
      6.4.1    Balance (optional)—Top loading, capable of weighing 500-2000 g within ± 1%
      6.4.2    Glass stirring rod.
      6.4.3    Separatory funnel—Glass, 2000-mL, with PTFE stopcock.
      6.4.4    Funnel—Large, glass, for pouring sample into separately funnel.
      6.4.5    Centrifuge (optional)—Explosion proof, capable of spinning at least  four 100-mL
              glass centrifuge tubes at 2400 rpm minimum.
      6.4.6    Centrifuge tubes (optional)—100-mL glass.
6.5    Equipment for removal of water, sodium sulfate, and silica gel fines.
      6.5.1    Funnel—Analytical, glass.
      6.5.2    Filter paper—Whatman No. 40 (or equivalent), to fit funnel.
6.6    Equipment for solvent distillation.
      6.6.1    Water bath or Steam bath—Explosion-proof, capable of maintaining a temperature
              of at least 85 °C.
      6.6.2    Flask—Boiling,  125-mL (Corning No. 4100 or equivalent).
      6.6.3    Distilling head—Claisen (VWR Scientific No.  26339-005, or equivalent), includes
              Claisen-type connecting tube and condenser.
      6.6.4    Distilling adaptor (attached to the distilling head and to the distillate collection flask
              for recovery of solvent).
      6.6.5    Distillate collection flask  (attached to  the distilling adaptor for  collection of the
              distilled solvent).
      6.6.6    Ice bath  or recirculating chiller  (to aid in the condensation and collection of the
              distilled solvent).
      6.6.7    Vacuum—Vacuum pump  or other source of vacuum.
      6.6.8    Tongs, for handling the boiling flask (Baxter Scientific Products No. T5007-2,  or
              equivalent).
      6.6.9    Desiccator—Cabinet- or jar-type, capable of keeping the boiling flask (Section 6.6.2)
              dry during cooling.
    6.6.10    Hood-Explosion-proof, capable of accommodating the equipment used for solvent
              distillation (Section 6.6.1-6.6.5).

6.7 Equipment for removal  of adsorbable  materials.

4                                                                              February 1999

-------
                                                                           Method 1664
      6.7.1    Magnetic stirrer.
      6.7.2    PTFE-coated magnetic stirring bars.
      6.7.3    Graduated cylinder—500-mL, capable of measuring ± 5 mL.
      6.7.4    Pipettes—Assorted sizes, calibrated to within ±0.5 percent.

7.0   Reagents and Standards
 7.1   Reagent water—Water in which HEM is not detected at or above the minimum level (ML)
       of this method.  Bottled distilled water or water prepared by passage of tap water through
       activated carbon have been shown to be acceptable sources of reagent water.
 7.2   Hydrochloric acid  or sulfuric acid—ACS.  Mix equal volumes of concentrated HC1 and
       reagent water or 1  part H2SO4 and 3 parts reagent water to produce an approximately 6N
       solution.
 7.3   n-Hexane—85% minimum purity, 99.0% min. saturated C6 isomers, residue less than 1 mg/L.
 7.4   Acetone—ACS, residue less than 1 mg/L.
 7.5   Sodium sulfate—ACS, granular anhydrous. Dry at 200-250 °C for 24 h minimum and store
       in a tightly sealed container until use.

    NOTE: Powdered sodium  sulfate should not be used because traces of
    water may cause it  to solidify.

 7.6   Boiling chips—Silicon carbide or fluoropolymer.
 7.7   Silica gel—Anhydrous, 75-150 micrometers, Davisil Grade 923 (Supelco 21447-7A, or
       equivalent). Dry at 200-250°C for 24 h minimum and store in a desiccator or tightly sealed
       container. Determine the n-hexane soluble material content of the silica gel by extracting 30
       g of silica gel with n-hexane and distilling the n-hexane to dryness.  The silica gel must contain
       less than 5 mg of n-hexane soluble material per 30 g (< 0.17 mg/g).
 7.8   Hexadecane—98% minimum purity.
 7.9   Stearic acid—98% minimum purity.
7.10   Hexadecane/stearic acid (1:1) spiking solution—Prepare in acetone at a concentration of 2
       mg/mL each.
     7.10.1    Place 200 ± 2 mg  stearic acid and 200 ± 2 mg hexadecane in a 100-mL volumetric
              flask and fill to the mark with acetone.

    NOTE: The solution  may require warming for complete dissolution of
    stearic  acid.

     7.10.2    After the hexadecane and stearic acid have dissolved,  transfer the solution  to a
              100-150 mL vial with fluoropolymer-lined cap. Mark the solution level on the vial
              and store in the dark at room temperature.
     7.10.3    Immediately prior to use, verify the level on the vial and bring to volume with
              acetone, if required.  Warm to redissolve all visible precipitate.
    NOTE: If  there is doubt of the  concentration, remove 10.0  +  0.1 mL with
    a volumetric pipet,  place in  a tared weighing pan,  and  evaporate to
    dryness in a fume  hood.  The  weight  must be  40 +  1 mg.  If not,  prepare
    a fresh solution  (Section 7.10.1).
 February 1999

-------
 Method 1664
7.11   Precision and  recovery (PAR) standard—Using a pipet, spike 10.0 ±0.1  mL of the
       hexadecane/stearic acid spiking solution (Section 7.10) into 950-1050 mL of reagent water
       to produce concentrations of approximately 20 mg/L each of hexadecane and stearic acid.
       The PAR standard is  used  for the determination of initial (Section  9.2.2) and  ongoing
       (Section 9.6) precision and recovery.
7.12   The spiking solutions should be checked frequently for signs of degradation or evaporation
       using the test noted in  Section 7.10.3, and must be replaced after six months, or sooner if
       degradation has occurred.

8.0   Sample Collection, Preservation,  and Storage
 8.1   Collect  approximately one  liter  of representative  sample in a glass bottle following
       conventional sampling practices (Reference 16.7), except that the bottle must not be pre-
       rinsed with sample before collection. To allow for potential QC failures, it is recommended
       that additional  sample aliquots be collected.
      8.1.1    If analysis is to be delayed for more than four hours, adjust the sample pH to less than
              2 with HC1 or H2SO4 solution (Section 7.2) at the time of collection, and refrigerate
              at 0-4 °C (40 CFR 136, Table II). To establish the volume of HC1 or H2SO4 required,
              collect  a separate aliquot, adjust the pH of this aliquot to less than 2 with acid, and
              add the volume of acid determined to each sample bottle prior to collection. Do not
              dip pH paper, a pH electrode, a stirring rod, or other materials into a sample that will
              be used for HEM or SGT-HEM determination because substances in the sample may
              adhere to these items.
      8.1.2   If a sample is known or suspected to contain greater than 500 mg/L of extractable
              material, collect a proportionately smaller volume of sample (the volume required will
              depend upon the estimated amount of extractable material) in a glass bottle. Add a
              proportionately smaller amount of HC1 or H2SO4 solution to the smaller sample if
              preservation is necessary.
 8.2   Collect an additional one or two aliquots (1L, additional smaller volume, or both) of a sample
       for each set of twenty samples or less for the matrix spike and, if used, the matrix spike
       duplicate.

    NOTE: For those  circumstances requiring the  collection of multiple
    aliquots  of one sample, each aliquot  is  to be collected  in either of
    the   following ways:  1)   collect simultaneously  in  parallel,  if
    possible,  or  2)  collect  as grab samples  in rapid  succession.

 8.3   The high probability that extractable matter may adhere to sampling equipment and result in
       measurements  that are biased low precludes the  collection of composite samples for
       determination of oil and grease. Therefore, samples must be collected as grab samples.  If a
       composite measurement is required, individual grab samples collected at prescribed time
       intervals must be analyzed separately and the concentrations averaged.  Alternatively, samples
       can be collected in the field and composited  in the  laboratory. For example, collect four
       individual 250-mL samples over the course  of a day. In the laboratory, pour each 250-mL
       sample into the separately funnel, rinse each of the four bottles (and caps) sequentially with
       30 mL of n-hexane, and use  the 30 mL of n-hexane for the extraction (Section 11.3).

6                                                                           February 1999

-------
                                                                               Method 1664
 8.4   All samples must be refrigerated at 0^°C from the time of collection until extraction (40
       CFR 136, Table II).
 8.5   All samples must be analyzed within 28 days of the date and time of collection (40 CFR 136,
       Table II).

9.0   Quality Control
 9.1    Each laboratory that uses this method is required to operate a formal quality assurance
       program (Reference 16.8). The minimum requirements of this program consist of an initial
       demonstration of laboratory capability, ongoing analyses of standards and blanks as a test of
       continued performance, and analysis of a matrix spike (MS) to assess recovery.  Laboratory
       performance is compared to established performance criteria  to determine if the results of
       analyses meet the performance characteristics of the method.
      9.1.1    The laboratory shall make an initial demonstration of the ability to generate acceptable
              accuracy and precision with this method.  This ability  is established as described in
              Section 9.2.
      9.1.2   In recognition of advances that are occurring in analytical technology, the laboratory
              is permitted certain  options to improve  separations or  lower the costs  of
              measurements, provided that all performance specifications are met. These options
              include alternate extraction and concentration devices and procedures such as solid-
              phase  extraction,   continuous   liquid-liquid  extraction,  and   Kuderna-Danish
              concentration.  Alternate determinative techniques, such as infrared spectroscopy or
              immunoassay, and changes that degrade method performance, are not allowed. If an
              analytical technique other than the techniques specified in this method is used, that
              technique must have  a  specificity equal to  or  better than the specificity  of the
              techniques  in this method for HEM and/or SGT-HEM in the sample  of interest.
              Specificity  is defined as producing results equivalent to the results  produced  by this
              method for  analytical standards (Section 9.2.2) and, where applicable, environmental
              samples (Section 9.2.3), and that meet all of the QC criteria stated  in this method.
           9.1.2.1    Each time a modification is made to this method, the laboratory is required to
                     repeat the IPR test in Section  9.2.2 to demonstrate that the modification
                     produces results equivalent to or superior to results produced by this method.
                     If the detection limit of the method will be affected by the modification, the
                     laboratory must demonstrate that the MDL (40 CFR Part 136, Appendix B)
                     is less than or equal to the MDL in this method or one-third the regulatory
                     compliance limit, whichever is higher.  If the modified method is to be used
                     for  compliance monitoring, the discharger/generator must also demonstrate
                     that the modified method recovers an amount of HEM and/or  SGT-HEM
                     equivalent  to  the  amount recovered by this method on each specific
                     discharge/waste stream. The tests required for this equivalency demonstration
                     are  given in Section 9.2.3.
           9.1.2.2   The laboratory is required to maintain records of modifications made to this
                     method. These records include the following, at a minimum:
                9.1.2.2.1     The names, titles, addresses, and telephone numbers of the analyst(s)
                            who performed the analyses  and modification, and of  the  quality

February 1999                                                                              7

-------
Method 1664
                            control  officer  who witnessed and  will verify the analyses  and
                            modification.
                9.1.2.2.2    A listing of pollutant(s) measured (HEM and/or SGT-HEM).
                9.1.2.2.3    A narrative stating reason(s) for the modification.
                9.1.2.2.4    Results from all quality control (QC) tests comparing the modified
                            method to this method, including:
                        (a) Calibration (Section 10).
                        (b) Calibration verification (Section 9.5).
                        (c ) Initial precision and recovery (Section 9.2.2).
                        (d) Analysis of blanks (Section 9.4).
                        (e) Accuracy assessment (Section 9.3).
                        (f) Ongoing precision and recovery (Section 9.6).
                        (g) Method detection limit (Section 9.2.1).
                9.1.2.2.5    Data that will  allow an  independent reviewer to  validate each
                            determination by tracing  the instrument output (weight or other
                            signal) to the final result.  These data are to include:
                        (a) Sample numbers and other identifiers.
                        (b) Extraction dates.
                        (c ) Analysis dates and times.
                        (d) Analysis sequence/run chronology.
                        (e) Sample weight or volume (Section 11.1.4).
                        (f) Extract volume for SGT-HEM (Section 11.5.2).
                        (g) Make and model of analytical balance and weights traceable toNIST.
                        (h) Copies of logbooks, printer tapes, and other recordings of raw data.
                        (i) Data system outputs, and other data to link the raw data to the results
                            reported.
      9.1.3    Analysis of a matrix spike (MS) is required to demonstrate recovery and to monitor
              matrix interferences (interferences caused by the sample matrix). The procedure and
              QC criteria for spiking are described in Section 9.3.
      9.1.4    Analyses  of laboratory blanks  are required  to demonstrate  freedom from
              contamination.  The procedure and criteria for analysis of a blank are described in
              Section 9.4.
      9.1.5    The laboratory shall, on an ongoing basis, demonstrate through calibration verification
              and analysis of the ongoing precision and recovery  (OPR) sample that the analysis
              system is  in control.   These procedures are described in Sections 9.5 and  9.6,
              respectively.
      9.1.6    The laboratory should maintain records to define the quality of data that is generated.
              Development of accuracy statements is described in Sections 9.3.7 and 9.6.3.
      9.1.7    Accompanying QC for the determination of HEM and/or SGT-HEM is required per
              analytical batch. For the definition of an analytical batch, see the glossary at the end
              of this method.
9.2    Initial demonstration of laboratory capability.
      9.2.1    Method Detection Limit (MDL)—To establish the ability to detect HEM and SGT-
              HEM, the laboratory shall determine the MDL per the  procedure in 40 CFR 136,

8                                                                             February 1999

-------
                                                                               Method 1664
              Appendix B using the apparatus, reagents, and standards that will be used in the
              practice of this method.  An MDL less than or equal to the MDL in Section 1.6 or less
              than 1/3 the regulatory compliance limit must be achieved prior to the practice of this
              method.
      9.2.1    Initial precision and recovery (IPR)—To establish the ability to generate acceptable
              precision and accuracy, the laboratory shall perform the following operations:
           9.2.2.1    Determine the concentration of HEM and/or SGT-HEM in four samples of
                     the PAR standard (Section 7.11) according to the procedure beginning in
                     Section 11.
           9.2.2.2    Using the results of the set of four analyses, compute the  average percent
                     recovery (X) and the standard deviation of the percent recovery (s) for HEM
                     and for SGT-HEM (if determined).  When determining SGT-HEM, the true
                     concentration (T) must  be  divided by 2 to reflect the concentration of
                     hexadecane that remains after removal of stearic acid. Use the following
                     equation for calculation of the standard deviation of the percent recovery:


                                       Equation 1
v -D
\
VY2 n(^X)
n
n -Dl
                          where:
                          n  =H\Number of samples
                          x  =D% Recovery in each sample
           9.2.2.3    Compare s and X with the corresponding limits for initial  precision and
                     recovery in Table 1.   If s and  X meet the acceptance  criteria,  system
                     performance is acceptable and analysis of samples may begin.  If, however, s
                     exceeds the precision limit or X falls outside the range for recovery, system
                     performance is unacceptable. In this event, correct the problem and repeat the
                     test.
      9.2.3   Equivalency demonstration for application of a method modification to compliance
              monitoring—To establish the ability of a modification of this method to recover an
              amount of HEM and/or SGT-HEM equivalent to the amount recovered  by this
              method from a specific discharge/waste stream, proceed as follows:
           9.2.3.1    Collect, extract, concentrate, and weigh the HEM or SGT-HEM in two sets
                     of four aliquots of unspiked wastewater.  One set of four wastewater aliquots
                     is analyzed according to the protocol in Section  11 of this method  and the
                     other set of four aliquots is analyzed using the modified  method.
           9.2.3.2    Calculate the average concentration of HEM and SGT-HEM for the set of
                     results from this method and for the set of results from the modified method.
                     The  average concentration using the modified method  must be 78 to 114
	percent of the average concentration produced by this method for HEM and
 February 1999                                                                            9

-------
Method 1664
                   64 to 132 percent of the average concentration produced by this method for
                   SGT-HEM. If not, the modified method may not be used.

   NOTE: If the average concentration of the four results produced using
   this  method  and  the  average   concentration  of the  four  results
   produced  using  the  modified method are  below the  minimum level
   (Section 1.6), and if the equivalency test of the modified method is
   passed for spikes  of reference standards into reagent water  (Section
   9.2.2) , the modified method is deemed to be equivalent to this method
   for determining HEM and or  SGT-HEM on that specific discharge/waste
   stream.

9.3   Matrix spikes—The laboratory must spike a minimum of 5 percent of all samples from a given
      sampling site or, if for compliance monitoring, from a given discharge/waste stream (matrix
      spike). The sample aliquot shall be spiked with the hexadecane/stearic acid spiking solution
      (Section 7.10). A duplicate matrix spike (MSD) is recommended but not required.
     9.3.1   The concentration of the spike in the sample shall be determined as follows:
         9.3.1.1    If, as in compliance monitoring, the concentration of HEM or SGT-HEM in
                   the sample is being checked against a regulatory concentration limit, the
                   spiking level  shall be at that limit,  at 1  to 5 times the  background
                   concentration of the  sample (determined  in  Section 9.3.2), or at the
                   concentration of the OPR (Section 9.4), whichever concentration is highest.
         9.3.1.2    If the concentration of HEM or SGT-HEM in a sample is not being checked
                   against a limit, the spike shall be at the concentration of the precision and
                   recovery  standard  (Section  7.11) or at  1  to  5 times higher than the
                   background concentration, whichever concentration is higher.
     9.3.2   Analyze one sample aliquot out of each set of 20 samples from each site or discharge/
            waste stream according to the procedure beginning in Section 11 to determine the
            background concentration (B) of HEM or SGT-HEM.
         9.3.2.1    If necessary, prepare a standard solution appropriate to produce a level in the
                   sample at the regulatory compliance limit or at 1 to 5 times the background
                   concentration (per Section 9.3.1).

   NOTE: Samples containing high concentrations (> 100 mg/L) of  HEM will
   require a large volume of spiking solution  (Section 7.10) for the MS
   (and  MSD).  If the concentration  of  HEM is expected  to exceed  1000
   mg/L,  smaller sample volumes should be collected for the background
   measurement  and MS  (and MSD)  so that  the amount  of  HEM  plus  the
   amount spiked does  not  exceed  1000  mg/L.

         9.3.2.2    Spike the additional sample aliquot(s) with the spiking solution and analyze
                   the aliquot(s) to determine the concentration after spiking (A).
     9.3.3   Calculate the percent recovery (P) of HEM or SGT-HEM in each aliquot using the
            following equation:
10                                                                    February 1999

-------
                                                                              Method 1664
                                       Equation 2

                                   p _D100 (A - B)
                                              TE
             where:
             AE=\]MeasuredEconcentrationEofEanalyteEafterEspikmg
             BE=\]MeasuredEbackgroundEconcentrationEofEHEMEorESGT-HEM
             TE= \]TrueEconcentrationEofEheEspike
             When determining SGT-HEM, the true concentration (T) must be divided by 2 to
             reflect the concentration of hexadecane that remains after removal of stearic acid.
     9.3.4   Compare the percent recovery of the HEM or SGT-HEM with the corresponding QC
             acceptance criteria in Table 1 .
          9.3.4.1    If the results of the spike fail the acceptance criteria, and the recovery of the
                    QC standard in the ongoing precision and recovery test (Section 9.6) for the
                    analytical batch is within the acceptance criteria in Table 1, an interference is
                    present.  In this case, the result may not be reported or used for purposes
                    regulatory compliance purposes and the laboratory must assess the potential
                    cause for the interference.  If the interference is attributable to  sampling, the
                    site or discharge/waste stream should be resampled.  If the interference is
                    attributable to a matrix problem, the laboratory must modify the method,
                    repeat the tests required in Section 9.1.2, and repeat the analysis of the sample
                    and the MS (and MSD, if performed). Most matrix interference problems are
                    attributable to the formation of emulsions in the extraction. Section 1 1.3.5
                    provides suggestions for overcoming emulsion problems.
          9.3.4.2   If the results of both the spike and the ongoing precision and recovery test fail
                    the acceptance criteria, the analytical system is judged to be out of control,
                    and the problem shall be identified and  corrected, and the  sample batch
                    reanalyzed.  All samples must be associated with a valid MS  (and MSD, if
                    performed).
     9.3.5   If an MSD was analyzed, compute the relative percent difference (RPD) between the
             MS and MSD (not between the two recoveries) using the following equation:


                                       Equation 3

                                        I D,-D,\
                              RPDE={]— - 1 - —   x  100
       where:
       Dl  =\^oncentrationEofEHEMEorESGT-HEMEnBheEsample
       D2  =\^oncentrationEofEHEMEorESGT-HEMEnBheEsecondE{duplicate}  sampleE
February 1999                                                                           11

-------
Method 1664
      9.3.6    The relative percent difference for duplicates shall meet the acceptance criteria in
              Table 1.  If the criteria are not met, the analytical system  is judged to be out of
              control, and the problem must be immediately identified and corrected, and  the
              analytical batch reanalyzed.
      9.3.7    As part of the QC program for the laboratory, it is suggested that method precision
              and accuracy for samples be assessed and records maintained. After the analysis of
              five spiked  samples in which the recovery passes the test in Section 9.3.4, compute
              the average percent recovery (Pa) and the standard deviation of the percent recovery
              (sp).  Express the accuracy assessment as a percent recovery interval from Pa - 2sp to
              Pa + 2sp. For example, if Pa = 90% and sp = 10% for five analyses of HEM or SGT-
              HEM,  the  accuracy interval is expressed as 70-110%.  Update  the  accuracy
              assessment  on a  regular  basis  (e.g., after each  five  to  ten new  accuracy
              measurements).
9.4    Laboratory blanks—Laboratory reagent water blanks are analyzed to demonstrate freedom
       from contamination.
      9.4.1    Extract and concentrate a laboratory reagent water blank initially (i.e. with the tests
              in Section 9.2) and with each analytical batch. The blank must be subjected to  the
              same procedural steps as a sample.
      9.4.2    If material is detected in the blank at a concentration greater than the minimum level
              (Section 1.6),  analysis of samples is halted until the source of contamination is
              eliminated and a blank shows no evidence of contamination. All samples must be
              associated with an uncontaminated method blank before the results may be reported
              for regulatory compliance purposes.
9.5    Calibration verification—Verify calibration of the balance per Section  10 before and after
       each analytical batch. If calibration is not verified after measurement of the analytical batch,
       recalibrate the balance and reweigh the batch.
9.6    Ongoing precision and recovery—To demonstrate that the analysis system is in control, and
       acceptable  precision and  accuracy  is being maintained  with  each analytical batch,  the
       laboratory shall perform the following operations:
      9.6.1    Extract and concentrate a precision and recovery standard (Section 7.11) with each
              analytical batch according to the procedure beginning in Section 11.
      9.6.2    Compare the recovery with the limits for ongoing precision and recovery in Table 1.
              If the recovery is in the range specified, the extraction, distillation, and weighing
              processes are in control and analysis of blanks and samples may proceed. If, however,
              the recovery is not in the specified range, the analytical process is not in control. In
              this event, correct the problem, re-extract the analytical batch, and repeat the ongoing
              precision and recovery test.
      9.6.3    The laboratory should add results that pass the specification in Section 9.6.2 to IPR
              and previous OPR data and update QC charts to form a graphic representation of
              continued laboratory performance.  The laboratory should also develop a statement
              of laboratory data quality for each analyte by calculating the average percent recovery
              (R) and the standard deviation of the percent recovery (sr). Express the accuracy as
              a recovery interval from R - 2sr to R + 2sr.  For example, if R = 95% and sr = 5%,  the
              accuracy is 85% to 105%.

12                                                                             February 1999

-------
                                                                           Method 1664
  9.7    Quality control sample (QCS)—It is suggested that the laboratory obtain a QCS from a
        source different from the source for the hexadecane and stearic acid used routinely in this
        method (Sections  7.8 and  7.9),  and that the QCS  be used for  verification of the
        concentrations of HEM and SGT-HEM using the procedure given in the note in Section
        7.10.3. The QCS should be analyzed monthly by laboratories performing routine analyses and
        less frequently by laboratories performing these analyses intermittently.

  9.8    The specifications contained in this method can be met if the apparatus used is scrupulously
        cleaned and dedicated for the determination of HEM and SGT-HEM. The standards used for
        initial precision and recovery (IPR, Section 9.2.2), matrix spike (MS, Section 9.3), and
        ongoing precision and recovery (OPR, Section 9.6) should be identical, so that the most
        precise results will be obtained.
  9.9    Depending upon specific program requirements, field  replicates and  field spikes of the
        analytes of interest into samples may be required to assess the precision and accuracy of the
        sampling and sample transporting techniques.

10.0    Calibration and Standardization
 10.1    Calibrate the analytical balance at 2 mg and 1000 mg using class "S" weights.
 10.2    Calibration shall be within ± 10% (i.e. ±0.2 mg) at 2 mg and ± 0.5% (i.e. ±5 mg) at 1000 mg.
        If values are not within these limits, recalibrate the balance.

11.0    Procedure
    This method is entirely empirical. Precise and accurate results can be obtained only by strict
    adherence to all details.

    NOTE: The procedure below is based  on the preparation,  extraction,E
    and analysis  of a 1-L  sample.   If a smaller volume is collected forE
    analysis, the laboratory should dilute the sample to 1  L with reagentE
    water  so  that  results  across  the  IPR,  blank,  OPR,  MS,  and,  ifE
    performed,  the MSD,  are  consistent.   It is also important  that  allE
    glassware surfaces be rinsed with n-hexane  to effect a quantitativeE
    transfer   of   the   constituents   in  the  sample   and   of   theE
    hexadecane/stearic acid in  the IPR,  OPR, MS,  and,  if performed,  theE
    MSD.E
 11.1    Preparati on of the analyti cal b atch.
      11.1.1   Bring the analytical batch of samples, including the sample aliquots for the MS (and
              MSD), to room temperature.
      11.1.1   Place approximately 1000 mL (950-1050 mL) of reagent water (Section 7.1) in a
              clean sample bottle to serve as the laboratory blank.
      11.1.3   Prepare the OPR (Section 9.6) using the PAR standard (Section 7.11).
      11.1.4   Either mark the sample bottle at the water meniscus or weigh the bottle for later
              determination of sample volume. Weighing will be more accurate.  Mark or weigh
              the MS (and MSD).
 11.2    pH verifi cati on.	
 February 1999                                                                       13

-------
 Method 1664
     11.2.1   Verify that the pH of the sample is less than 2 using the following procedure:
          11.2.1.1    Dip a glass stirring rod into the well mixed sample.
          11.2.1.2    Withdraw the stirring rod and allow a drop of the sample to fall on or touch
                    the pH paper.

    NOTE: Do  not  dip the pH paper  into  the  bottle or  touch it  to theE
    sample on the  lid.E

          11.2.1.3    Rinse the stirring rod with a small portion of n-hexane that will be used for
                    extraction (to ensure that no extractable material is lost on the stirring rod).
                    Collect the rinsate in the separatory funnel to be used for sample extraction.
     11.2.2   If the sample is at neutral pH, add 5-6 mL of HC1 or H2SO4 solution (Section 7.2) to
             the 1-L sample.  If the sample is at high pH, use a proportionately larger amount of
             HC1  or H2SO4 solution.   If a smaller  sample  volume was collected, use  a
             proportionately smaller amount of HC1 or H2SO4 solution.
     11.2.3   Replace the cap and shake the bottle to mix thoroughly. Check the pH of the sample
             using the procedure in Section 11.2.1. If necessary, add more acid to the sample and
             retest.
     11.2.4   Add the appropriate amount of HC1 or H2SO4 solution to the blank, OPR, MS (and
             MSD) to adjust the pH of these solutions to <2.

    NOTE: The  procedure detailed below is for separatory funnel liquid-
    liquid extraction.  Solid-phase extraction  (SPE)  may be used at theE
    discretion of the discharger/generator and its laboratory.  However,E
    if SPE is  used, it  is  the responsibility of the discharger/generatorE
    and  laboratory to  assure that results  produced  are  equivalent  toE
    results produced  by  the  procedure below.E

11.3   Extraction
     11.3.1   Tare a clean boiling flask containing 3-5 boiling chips as follows:
          11.3.1.1    Place the flask containing the chips in an oven at 105-115 °C for a minimum
                    of 2 h to dry the flask and chips.
          11.3.1.2    Remove from the oven and immediately transfer to a desiccator to cool to
                    room temperature.
          11.3.1.3    When cool, remove from the desiccator with tongs and weigh immediately on
                    a calibrated balance  (Section 10).
     11.3.2   Pour the sample into the separatory funnel.
     11.3.3   Add 30 mL  of n-hexane to the sample bottle and seal the bottle with the original
             bottle cap. Shake the bottle to rinse all interior surfaces of the bottle, including the
             lid of the bottle cap.  Pour the solvent into the separatory funnel.
     11.3.4   Extract the sample by shaking the separatory funnel vigorously for 2 minutes with
             periodic venting into a hood to release excess pressure.
     11.3.5   Allow the organic phase to  separate from the aqueous phase for a minimum of 10
             minutes.  If an emulsion forms between the phases and the emulsion is greater than
             one-third the volume of the solvent layer, the laboratory must employ emulsion-
             breaking techniques to complete the  phase separation.  The optimum technique
             depends upon the sample, but may include stirring, filtration through glass wool, use

 14                                                                        February 1999

-------
                                                                         Method 1664
             of solvent phase separation paper, centrifugation, use of an ultrasonic bath with ice,
             addition of NaCl, or other physical methods. Alternatively, solid-phase extraction
             (SPE), continuous liquid-liquid extraction, or other extraction techniques may be used
             to prevent emulsion formation, provided that the requirements in Section 9.1.2 are
             met.
    11.3.6    Drain the aqueous layer (lower layer) into the original sample container. Drain a small
             amount of the organic layer into the sample container to minimize the amount of
             water remaining in the separatory funnel.

   NOTE: The  amount  of  water  remaining  with  the  n-hexane  must beE
   minimized to prevent dissolution or clumping  of the  sodium  sulfateE
   in  the solution drying process. E

    11.3.7    Place a filter paper (Section 6.5.2) in a filter funnel (Section 6.5.1), add approximately
             10 g of anhydrous Na2SO4, and rinse with a small portion of n-hexane. Discard the
             rinsate.

   NOTE: The specific properties of  a sample may necessitate the use ofE
   larger amounts  of Na2SO4.E

    11.3.8    Drain the n-hexane layer (upper layer) from the separatory funnel through the Na2SO4
             into the pre-weighed boiling flask containing the boiling chips (Section 11.3.1.3).

   NOTE: It  is  important  that water  be removed  in this  step.   WaterE
   allowed to  filter through the Na2SO4 will dissolve some of the Na2SO4g
   and carry it into the boiling flask compromising the  determination.E

    11.3.9    Repeat the extraction (Sections 11.3.3-11.3.6and 11.3.8) twice more with fresh 30-
             mL portions of n-hexane, combining the extracts in the boiling flask.
   11.3.10    Rinse the tip of the separatory funnel, the filter paper, and the funnel with 2-3  small
             (3-5 mL) portions  of n-hexane.  Collect the rinsings in the flask.

   NOTE: For samples that  are expected to  contain a high concentrations
   of  salt  (e.g.,  waters  from oil  production  facilities), it  may beE
   prudent  to collect the extract  in a  250-mL separatory funnel andE
   back-extract with reagent water.  After back-extraction,  the  extracts
   should be drained through  Na2SO4 to remove  all traces  of water.E

   11.3.11    A milky extract  indicates the presence of water. If the extract is milky, allow the
             solution to stand for up to one hour to allow the water to settle.  Decant the solvent
             layer (upper layer) through sodium sulfate to remove any excess water as in Sections
             11.3.7  and 11.3.8.  Rinse the glassware and sodium sulfate with small portions of n-
             hexane to effect  a quantitative transfer.
   11.3.12    If only SGT-HEM is to be determined, proceed to Section 11.5.

   11.4   Solvent distillation.
    11.4.1    Connect the boiling flask to the distilling head apparatus and distill the solvent by
             immersing the lower half of the flask in a water bath or a  steam bath. Adjust the

February 1999                                                                      15

-------
 Method 1664
              water temperature as required to complete the concentration in less than 30 minutes.
              Collect the solvent for reuse.
     11.4.2    When the temperature in the distilling head reaches approximately 70°C or the flask
              appears almost dry, remove the distilling head.  Sweep out the flask for 15 seconds
              with air to remove solvent vapor by inserting a glass tube connected to a vacuum
              source. Using tongs, immediately remove the flask from the heat source and wipe the
              outside surface dry to remove moisture and fingerprints.

    NOTE: The laboratory should carefully monitor the flask during the
    final stages of distillation  to assure  that all of  the solvent  is
    removed and to prevent loss of the more volatile sample constituents.

     11.4.3    Inspect the residue in the boiling flask for crystals. Crystal formation is an indication
              that sodium sulfate may have dissolved and passed into the boiling flask.  This may
              happen if the drying capacity of the sodium sulfate is exceeded or if the sample is not
              adjusted to low pH.  If crystals are observed, redissolve the extract in n-hexane,
              quantitatively transfer through a filter into another tared boiling flask, and repeat the
              distillation procedure (Sections  11.4.1-11.4.2).
     11.4.4    Dry the boiling flask for 30 - 45 minutes in an oven maintained at 70 ± 2 °C.  Cool
              to room temperature in a desiccator and maintain in the desiccator for 30 minutes
              minimum. Remove with tongs and weigh immediately.  Repeat the cycle of drying,
              cooling, desiccating, and weighing until the weight  loss is less than 4 %  of the
              previous weight or less than 0.5  mg, whichever is less.
          11.4.4.1   If the extract was from the HEM procedure,  determine the HEM (Wh) by
                     subtracting the tare weight (Section 11.3.1) from the total weight of the flask.
          11.4.4.2   If the extract was from the SGT-HEM procedure (Section 11.5.5), determine
                    the weight of SGT-HEM (Ws) by subtracting the tare weight from the total
                    weight of the flask.
     11.4.5    Determine the original sample volume (Vs) in liters by  filling the sample bottle to the
              mark with water and measuring the volume of water in a 1- to 2-L graduated cylinder.
              If the sample weight was used (Section 11.1.4), weigh the empty bottle and cap and
              determine Vs by difference, assuming a sample density of 1.00.
11.5   SGT-HEM determination.
     11.5.1    Silica gel  capacity—To ensure that the capacity of the silica gel will not be exceeded,
              the amount of HEM must be less than 100 mg or, if above 100 mg, must be known.
          11.5.1.1   If it is known that the amount of HEM is less than 100 mg, the laboratory may
                    proceed with the determination of SGT-HEM per Sections 11.5.3-11.5.5
                    without determination of HEM.
          11.5.1.2   If, however, the amount of HEM is not known, HEM must first be determined
                    using the procedure in Sections 11.3-11.4.
     11.5.3    Extractable materials in silica gel—Because the capacity of silica gel is not known for
              all substances, it is presumed that 3 g will normally adsorb  100 mg of all adsorbable
              materials. Therefore, for samples containing 1000 mg HEM, 30 g of silica gel will be
              needed.  The amount of silica gel that can be used for adsorption in the SGT-HEM
              procedure below has been limited  to 30  g because of concerns about possible

 16                                                                          February 1999

-------
                                                                                Method 1664
               extractable impurities in the silica gel.  If the amount of HEM in the sample is greater
               than 1000 mg, split the extract per the following procedure:
           11.5.2.1    Add  85-90 mL of n-hexane to the boiling flask to redissolve the HEM.  If
                      necessary, warm the solution to completely redissolve the HEM.
           11.5.2.2    Quantitatively transfer the extract to a 100-mL volumetric flask. Dilute to the
                      mark with n-hexane.
           11.5.2.3    Calculate the extract volume that contains 1000 mg of extractable material
                      according to the following  equation:


                                        Equation 4

                                             ,1000 V.
                                       V =D-
                                                w,.
                                                 h
                where:
                Va  -Volume of aliquot to be withdrawn (mL)
                 Vt  -\\Total volume of solvent used in Section 11.5.2.2 (mL)
                Wh  =H\Weight of extractable material HEM measurement (mg)
           11.5.2.4    Using a calibrated pipet, remove the volume to be withdrawn (Va) and return
                      to the boiling flask. Dilute to approximately 100 mL with n-hexane.
      11.5.3   Adsorption with silica gel
           11.5.3.1    Add 3.0 ± 0.3 g of anhydrous silica gel (Section 7.7) to the boiling flask for
                      every 100 mg of HEM, or fraction thereof, to a maximum of 30 g of silica gel.
                      For example, if the weight of HEM is 735 mg, add 3 x 8 = 24 g of silica gel.
           11.5.3.2    Add a fluoropolymer-coated stirring bar to the flask and stir the solution on
                      a magnetic stirrer for a minimum of 5 minutes.
      11.5.4   Filter the solution through n-hexane moistened filter paper into a pre-dried, tared
               boiling flask containing several boiling chips. Rinse the silica gel and filter paper with
               several small amounts of n-hexane to complete the transfer.
      11.5.5   Distill the solution and determine the weight of SGT-HEM per Section 11.4.

12.0   Data Analysis and  Calculations
 12.1   n-Hexane extractable material—Calculate the concentration of HEM ("oil and grease") in the
        sample per the following  equation:
  February 1999                                                                           17

-------
  Method 1664
                                       Equation 5

                                                 W, (mg)
                                 HEM (mg/L) =D  h

              where:
              Wh =UWeight of extractable material from Section 11.4.4.1  (mg)
              V  =USample volume from Section 11.4.5  (L)
 12.2   Silica gel treated n-hexane extractable material—Calculate the concentration of SGT-HEM
        ("non-polar material") in the sample per the equation above, substituting Ws (from Section
        11.4.4.2) for Wh. If the extract was split to decrease the total amount of material to 1,000
        mg, determine the corrected total weight of SGT-HEM in the un-split extract (Wc) using the
 	following equation:	

                                       Equation 6


                                   Wc (mg)  =D-i Wd (mg)
                                               V
                                               a
 where:
  Wd =UWeight in the portion  of the extract split for adsorption (Sections 11.5.2.4 and 11.4.4.2)
  Vt  and Va are as  defined in Equation 4
        Use the corrected total weight of SGT-HEM in the unsplit extract (Wc) to determine the total
        SGT-HEM in the sample by substituting Wc for Wh in Equation 5.
 12.3   Reporting—Report results to three significant figures for HEM and SGT-HEM found at or
        above 10 mg/L, and report results to two significant figures for HEM and SGT-HEM found
        below 10 mg/L.
      12.3.1    Samples—Report results for HEM and SGT-HEM found below the ML as < 5.0
               mg/L, or as required by the permitting authority or permit.
      12.3.2    Blanks—Report results for HEM and SGT-HEM  found below the MDL as < 1.4
               mg/L, or as required by the permitting authority or permit.  Do not report results
               below the MDL unless required by the permitting authority or permit.
      12.3.3    Results from tests performed with an analytical system that is not in control (Section
               9) must not be reported or otherwise used for permitting or regulatory compliance
               purposes but do not relieve a discharger or permittee of timely reporting.

13.0   Method Performance
 13.1   This method was validated through single laboratory studies and an inter-laboratory method
        validation study (Reference 16.9).   Combined data from these studies show an average
  18                                                                          February 1999

-------
                                                                              Method 1664
        recovery of 93 percent for HEM and 89 percent for SGT-HEM and precision (as relative
        standard deviation) of 8.7 percent for HEM and 13 percent for SGT-HEM.
 13.2   The method detection limit (MDL) and minimum level of quantitation (ML) are based on five
        studies conducted by EPA and described at proposal of Method 1664 (61 FR 1730) and as
        verified by data submitted in comments on the proposal of Method 1664 (Reference 16.3).

14.0   Pollution Prevention
 14.1   The solvents used in this method pose little threat to the environment when recycled and
        managed properly.
 14.2   Standards  should be prepared in volumes consistent with laboratory use to minimize the
        volume of expired standards to be disposed.

15.0   Waste Management
 15.1   It is the laboratory's responsibility to comply with all Federal, State, and local regulations
        governing  waste management, particularly the hazardous waste identification rules and land
        disposal restrictions, and to protect the air, water, and land by minimizing and controlling all
        releases from fume hoods and bench operations.  Compliance  with all sewage  discharge
        permits and regulations is also required.
 15.2   Samples preserved with HC1  or H2SO4 to pH < 2 are hazardous and must be neutralized
        before being disposed, or must be handled as hazardous waste.
 15.3   For further information on waste management, consult "The Waste Management Manual for
        Laboratory Personnel," and "Less is Better:  Laboratory Chemical Management for Waste
        Reduction," both available from the American Chemical Society's Department of Government
        Relations and Science Policy, 1155 16th Street N.W., Washington, D.C. 20036.

16.0   References
 16.1      "Methods for Chemical Analysis of Water and Wastes," 3rd Edition, Environmental
           Protection Agency, Environmental Monitoring Systems Laboratory-Cincinnati (EMSL-
           Ci), Cincinnati, Ohio 45268, EPA-600/4-79-020, Method 413.1, (1983).
 16.2      Ibid., Method 418.1
 16.3      Guidelines Establishing Test Procedures for the Analysis of Oil and Grease and Non-polar
           Materials; Final Rule; Preamble, Responses to Comments, and Docket, as referenced in
           the Final Rule.
 16.4      "Carcinogens -  Working  With  Carcinogens," Department of Health, Education, and
           Welfare, Public  Health Service,  Center for Disease Control, National Institute for
           Occupational Safety and Health, Publication No. 77-206, August 1977.
 16.5      "OSHA Safety and Health Standards, General Industry," (29 CFR 1910), Occupational
           Safety  and Health Administration, OSHA 2206 (Revised, January 1976).
 16.6      "Safety in Academic Chemistry Laboratories," American Chemical Society, Committee
           on Chemical Safety, 3rd Edition, 1979.
 16.7      "Standard Practices for Sampling Water," ASTM Annual Book of Standards, Part 31,
           D3370-76, American Society for Testing and Materials, 1916 Race Street, Philadelphia,
           PA 19103-1187, 1980.

 February 1999                                                                          19

-------
  Method 1664
 16.8      "Handbook of Analytical Quality Control in Water and Wastewater Laboratories,"
           USEPA, EMSL-Ci, Cincinnati, OH 45268,  EPA-600/4-79-019, March 1979.
 16.9      Report of the Method 1664 Validation Studies, April 1995. Available from the Sample
           Control Center (operated by DynCorp I&ET), 6101 Stevenson Avenue, Alexandria, VA
           22304,(703)461-2100.

17.0   Tables
 Table 1. Acceptance Criteria for Performance Tests


    Acceptance Criterion            Section              Limit (%)


    Initial precision and recovery              9.2.2
        HEM Precision (s)                  9.2.2.2                       11
        HEM Recovery (X)                  9.2.2.2                   83-101
        SGT-HEM Precision (s)              9.2.2.2                       28
        SGT-HEM Recovery (X)             9.2.2.2                   83-116


    Matrix spike/matrix spike duplicate          9.3
        HEM Recovery                     9.3.4                   78-114
        HEMRPD                         9.3.5                       18
        SGT-HEM Recovery                 9.3.4                   64-132
        SGT-HEM RPD                     9.3.5                       34


    Ongoing precision and recovery            9.6
        HEM Recovery                      9.6                   78-114
        SGT-HEM Recovery                  9.6                   64-132


18.0   Glossary of Definitions and Purposes
        The definitions and purposes are specific to this method but have been conformed to common
        usage to the extent possible.
 18.1   Units of weight and measure and their abbreviations
      18.1.1    Symbols
           °C        degrees Celsius
           <        less than
           %        percent
           ±        plus or minus
      18.1.2    Alphabetical characters
 	g	gram	
 20                                                                        February 1999

-------
                                                                                Method 1664
           h         hour
           L         liter
           mg       milligram
           mg/g      milligram per gram
           mg/L      milligram per liter
           mg/mL    milligram per milliliter
           mL       milliliter
           No.           number
           rpm       revolutions per minute
18.2   Definitions, acronyms, and abbreviations
     18.2.1   Analvte: The HEM or SGT-HEM determined by this method.
     18.2.2   Analytical batch: The set of samples started through the extraction process in a 12-
              hour shift, to a maximum of 20 field samples. Each analytical batch of 20 or fewer
              samples must be accompanied by a laboratory  blank  (Section 9.4), an ongoing
              precision and recovery sample (OPR, Section 9.6), and a matrix spike, (Section 9.3),
              resulting in a minimum of four analyses (1 sample, 1 blank, 1 OPR, and 1 MS) and a
              maximum of 23 analyses (20 field samples, 1 blank,  1  OPR, and 1 MS) in the batch.
              If greater than 20 samples are to be extracted in a 12-hour shift, the samples must be
              separated into analytical batches of 20 or fewer samples.
     18.2.3   Discharge (matrix type): A sample medium with common characteristics across a
              given industrial subcategory (40 CFR parts 403-500). For example, C-stage effluents
              from chlorine bleach mills in the Pulp, Paper, and Paperboard industrial category;
              effluent from the Continuous Casting subcategory of the Iron and Steel industrial
              category; publicly owned treatment work (POTW) sludge; and in-process streams in
              the Atlantic and Gulf Coast Hand-shucked Oyster Processing subcategory are each
              a matrix type.
     18.2.4   Field blank: An aliquot of reagent water that is placed in a sample container in the
              laboratory or in the field and treated as a sample in  all respects, including exposure
              to sampling site conditions, storage, preservation, and all analytical procedures. The
              purpose of the field blank is to determine if the field or sample transporting
              procedures and environments have contaminated the sample.
     18.2.5   HEM: See n-Hexane extractable material.
     18.2.6   n-Hexane extractable material:  Material  that is  extracted from  a sample and
              determined by this method (oil and grease).  This material includes  relatively non-
              volatile hydrocarbons, vegetable oils, animal fats, waxes, soaps, greases, and related
              matter.
     18.2.7   IPR:  See initial precision and recovery.
     18.2.8   Initial precision and recovery (IPR): Four aliquots  of the diluted PAR analyzed to
              establish the ability  to generate acceptable precision and accuracy.  An  IPR is
              performed the first time this method is used and any time the method is modified.
     18.2.9   Laboratory blank (method blank): An ali quot of reagent water that i s treated exactly
              as a  sample  including exposure to all glassware,  equipment,  solvents, reagents,
              internal standards, and surrogates that are used with  samples.  The laboratory blank
              is  used to determine if analytes or interferences  are present in  the laboratory
              environment, the reagents, or the apparatus.

 February 1999                                                                            21

-------
Method 1664
   18.2.10    Laboratory control sample (LCS):  See Ongoing precision and recovery standard
              (OPR).
   18.2.11    Matrix spike (MS) and matrix spike duplicate (MSP): Aliquots of an environmental
              sample to which known quantities of the analytes are added in the laboratory.  The
              MS and MSD are prepared and/or analyzed exactly like a field sample. Their purpose
              is to quantify any additional bias and imprecision caused by the sample matrix.  The
              background concentrations of the analytes in the sample matrix must be determined
              in a separate aliquot and the measured values in the MS and MSD corrected for
              background concentrations.
   18.2.12    May:  This action, activity, or procedural step is neither required nor prohibited.
   18.2.13    May not: This action,  activity,  or procedural step is prohibited.
   18.2.14    Method Detection Limit: The lowest level at which an analyte can be detected with
              99 percent confidence  that the analyte concentration is greater than zero.
   18.2.15    Minimum Level (ML): The lowest level at which the entire analytical system gives
              a recognizable signal and acceptable calibration point for the analyte. It is equivalent
              to the concentration of the lowest calibration standard, assuming that all method-
              specified sample weights, volumes, and cleanup procedures have been employed.
   18.2.16    Must: This action, activity, or procedural step is required.
   18.2.17    Ongoing precision and recovery standard (OPR. also called a  laboratory control
              sample): A laboratory blank spiked with known quantities of analytes. The OPR is
              analyzed exactly like a sample.  Its purpose is to assure that the results produced by
              the laboratory remain  within the limits  specified  in this method for precision and
              accuracy.
   18.2.18    OPR:  See ongoing precision and recovery standard.
   18.2.19    PAR:  See precision and  recovery standard.
   18.2.20    Precision and recovery standard: Secondary standard that is diluted and spiked to
              form the IPR and OPR.
   18.2.21    Quality control sample (PCS):   A sample containing analytes of interest at known
              concentrations.  The QCS is obtained from a source external to the laboratory or is
              prepared  from standards obtained from a  different source than the calibration
              standards. The purpose is to check laboratory performance using test materials that
              have been prepared independently from  the normal preparation process.
   18.2.22    Quantitative transfer:   The process of transferring a solution from one container to
              another using a pipet in which as much solution as possible is transferred, followed
              by rinsing of the walls of the source container with a small volume of rinsing solution
              (e.g., n-hexane), followed by transfer of the rinsing solution, followed by a second and
              third rinse and transfer.
   18.2.23    Reagent water:  Water demonstrated to be free  from HEM and SGT-HEM and
              potentially interfering substances at or above the minimum level of this method.
   18.2.24    Regulatory Compliance Limit: A limit on the concentration or amount of a pollutant
              or contaminant specified in a nationwide  standard, in  a permit, or  otherwise
              established by a regulatory authority.
   18.2.25    SGT-HEM:  See Silica gel treated n-hexane extractable material.
   18.2.26    Should:  This action, activity, or procedural step is suggested but not required.
22                                                                            February 1999

-------
                                                                                Method 1664
   18.2.27    Silica gel treated n-hexane extractable material:  Components of n-Hexane extractable
              material (HEM) that are not adsorbed by silica gel; i.e., non-polar material (NPM).
   18.2.28    Stock solution:  A solution containing an analyte that is prepared using a reference
              material traceable to EPA, the National Institute of Science and Technology (NIST),
              or a source that will attest to the purity and authenticity of the reference material.
February 1999                                                                             23

-------