United States          September
Environmental Protection     1986
Agency


        -

Design,
Construct, and
Operate for



-------
Disinfection  with  Ultraviolet
Introduction
The use of ultraviolet radiation for the disinfection of
wastewater is an effective and economical alternative
to chlorination and ozonation. Ultraviolet (UV)
disinfection, like chlorination, involves the addition of a
germicidal agent to a wastewater to inactivate the
bacteria. Like ozonation, disinfection  with UV radiation
requires on-site generation of the germicidal agent. The
principal advantage of UV disinfection over chlorination
is that UV leaves no residual in the wastewater that
may affect the receiving waters.

As shown in Table 1, there are a number of UV
installations currently operating in the U.S. The list
reflects the distribution of treatment plants with regard
to size. The operating plants are predominantly small,
although those in the planning or construction stages
are larger.
Size (Design)
(m3/day) (MGD)
<380 (<0.1)
380-1900 (0.1-0.5)
1900-3800 (0.1-1.0)
3800-19000(1.5)
19000-38000 (5-10)
In
Operation
15
17
7
11
-
Under
Construction
—
10
5
14
-
38000-190000 (10-50) 1
>1 90000 (>50)
Total
Note: List compiled
-
51
Spring 1984.
1
30

Being
Designed
7
10
4
11
-
2
-
34

Table 1. Summary of UV Installations in the United States.
The Process
Disinfection by UV radiation relies on the transference
of ultraviolet electromagnetic energy from a lamp
source to an organism's cellular material to prevent cell
replication. The effectiveness of the radiation depends
on the dose and the organism's exposure time.
The most efficient and effective artificial source of UV
energy is the mercury lamp. It can be submerged in or
suspended above the wastewater. If submerged, it is
inserted into a quartz sleeve to minimize the cooling
effects of the water. Figure 1 is a schematic
representation of a UV disinfection unit showing a
submerged lamp placed perpendicular to the direction
of the wastewater flow. Lamps may also be placed

-------
    The estimation of K and Np involves generating a set
    of data specifically for this purpose during the design,.
    phase of the project. The rate of bacterial inactivation
    (K) is a function of intensity as shown on Figure 2.
    Increasing the intensity of the UV radiation increases  "
    the rate,  K. The coliform density associated with
    particulates, Np, is a function of the suspended solids in
    the wastewater.  When these relationships are
    established, the performance of the UV system can be
    approximated for different equipment configurations and
    process variations.

    Design
    The degree of pretreatment received before the       !
    disinfection step affects the sizing and performance of
    the UV system. The five wastewater parameters that
    most affect UV design and performance are flow (Q),
    suspended solids (ss), initial bacteria density (N0),
    density of bacteria associted with particulates (Np), and
    the UV absorbence of the wastewater.                r

    The performance of a UV disinfection system is directly
    related to the initial  density (N0) of the indicator
    organisms . Performance is measured as the log of the;
    survival  ratio, N/N0  (see Figure 2). The initial density
    cannot be predicted based on the type of pretreatment
    received and should be measured prior to design.

    The  aggregation of  bacteria and particulates,  expressed
    as Np, significantly affects the efficiency of UV
    disinfection. The level of Np is related to the suspended
    solids content.                                      !
                              alt,
                          watts/cm)
     to the initial density (N0) of the indicator organisms.
I for a given residence time.

-------
The UV "demand" of the chemical constituents of the
wastewater must also be measured since this
absorbence will affect the intensity of the radiation
within the reactor. In specific design situations, the level
of absorbence will affect the sizing of a system and
possibly the configuration of the lamps.

Design Considerations
• Low pressure mercury arc lamps are currently the
  most efficient source of UV radiation.

• Temperature effects should be minimized. In quartz
  systems, O-ring spacers should be slipped over the
  lamps to prevent direct contact with the cooler quartz
  sleeve.
• Fittings  holding the quartz sleeve should be tight and
  leakproof to prevent fouling on the inside of the
  sleeve.
• The control panel should  be remote from the UV
  reactor.
• The ballasts must be properly mated with the lamps
  and thermally protected to shut down if they
  overheat.
• The electrical wiring should be properly sized and be
  resistant to UV radiation effects.
• Large debris should be prevented from entering the
  UV system.
• A feature that has been found not  to be successfully
  applied is the use of microprocessor controlled
  automatic lamp bank shut-off systems. Many plants
  that originally had automatic systems have
  since reverted to manual  control. The use of a simple
  mechanical cam timer that shuts off selected lamps
  or banks of lamps to reduce UV output during
  selected time periods is recommended.
Operational Considerations

• Cleaning
  Periodic chemical or detergent cleaning of surfaces
  that come in contact with the wastewater is required,
  particularly when the wastewater has a high oil and
  grease content or has a high hardness content.
  Magnesium and calcium carbonate deposits are a
  major cause of fouling of quartz surfaces. However,
  acidification of the reactor water will usually restore
  the surface. Organic fouling  by oil and grease must

-------
rate  for  Success
        N = (N'0 + Np) exp (-Kit) + Nf
                                             (2)
        where:    N'0 = initial, non-aggregate density
                         (conforms per 100 ml)
                  Np = coliform density associated with
                        particulates (coliforms per 100 ml)
    Generally N'0»NP,  and N0 = N0' + Np; therefore.
    Equation 2 can be rewritten as Equation 3.
        N = N0 exp (-Kit) + Np
                                             (3)
    Equation 4 incorporates the dispersive properties of a
    reactor under steady-state conditions.
N = N0 exp
                                      N
(4)
       where:     z =  1  + 4KE/U2
                  x =  length of reactor, cm
                  E =  dispersion coefficient, cm2/sec
                  K =  rate of bacteria I inactivation, sec;"1
                  u =  velocity of wastewater, cm/sec '
       where:     u =  xVv"1CT1
                  Vv = void volume of reactor,  liters
                  Q =  wastewater flow, liters/sec
                                    I  = Intensity
                               I       I1>I2>I3
                                      Time
    Figure 2. The performance of a UV disinfection system as
            Note: The rate K increases with increasing intensil

-------
 be cleaned with detergents or a combination of
 cleaning methods.

 Accessory equipment is available to maintain the
 surfaces. These include the mechanical wiper,
 ultrasonic transducer, or a higher pressure spray
 nozzle. While these methods are somewhat effective,
 intermittent cleaning with chemicals is generally
 required.
• Elements for Effective Maintenance
 The reactor should be designed with drains that
 allow complete and rapid emptying.
 The system should be modularly designed to allow
 isolation of any unit from the  plant flow.
 Strict inventories of lamp use and output should be
 maintained.
 The reactor should be drained when removed from""
 service.
> Labor
 The labor requirements  of the UV system are divided
 into three main categories: direct UV operation and
 maintenance tasks, general maintenance, and
 system overhaul. Overall, the total labor needs for the
 UV process are relatively low, ranging from
 approximately 40 persondays/year for a small  10  KW
      1000
                 Approximate Number of Lamps (1.5 ARC)
                 60   120   300  600  1200   3000 6000
                                   Note:
                                   Labor Based on
                                   Total System KW
                             40 60 100  200 4006001000
Figure 3. Estimate of Labor Requirements for the
         Operation and Maintenance of UV Systems
             Scheible, O. K., et. al.,  1986.

-------
|ht  -   Design,  Construct,  and  Ope
    Influent
                                         Control Box
                                          UV Lamp Inside
                                          Quartz Sleeve
                                              Effluent
   Figure 1.  Ultraviolet Design Schematic.

   parallel to the flow in some UV units. The intensity of
   the radiation emitted by the lamp dissipates as the
   distance from the lamp increases.

   Process  Model - UV Reactor Performance
   A mathematical model was developed during a study2
   at Port Richmond, NY to account for major process
   variables  in the design of a UV system. The derivation
   of this expression is briefly presented here. Inactivation
   of bacteria by  UV radiation can be approximated by
   the following equation:
      N = N0 exp (-Kit)
0)
     where:   N = effluent bacterial density (coliforms per
                  100 ml)
              N0 = influent bacterial density (coliforms
                   per 100 ml)
              K = rate constant (cm2watT1sec~1)
              I = intensity watt(cm2)"1                 i
              t = time of exposure (sec)               ;;
   Equation 1 is a good first approximation of a response"
   to a given dose. However, direct testing on mixed     |
   cultures often shows a reduced efficiency with
   increasing dose. In wastewater treatment, this has
   been attributed to the aggregation of bacteria and
   particulate matter.  Ultraviolet light is unable to penetrate
   this material  to inactivate the bacteria; thus, the
   continued elevation of dose will show a diminishing
   response as residual active bacteria are protected in   '
   the particulates. Consequently, Equation  1 should be
   modified (shown as Equation 2) to account for the
   effect of particulate matter in the wastewater.

-------
  (120 lamps) system to approximately 40
  persondays/year for a 400 KW (5000 lamps) system.
  Total yearly estimated labor requirements are
  presented in Figure 3.
Summary
Studies have found UV to be very effective for
disinfection of secondary or tertiary effluent. In addition,
the studies showed quartz systems to be generally
more energy  efficient than teflon systems.
Also, the efficiency of both systems drop significantly
as turbulence and suspended solids  levels increase
above secondary levels.
The UV process is relatively simple and also  offers the
advantages of system flexibility and capability of
responding to changes in demand. Since the process
leaves no residual in  the wastewater that could impact
the receiving water, less rigorous control is necessary
than  that associated with chlorine.  In addition, a
mathematical model recently developed will aid in the
design of future UV disinfection  systems.

Additional Reading
1. U.S. Environmental Protection Agency, Technology Transfer
  Process Design Manual for Municipal Wastewater Disinfection,
  U.S. Environmental Protection Agency, CERI, Cincinnati, OH, 1986.
2. Scheible, O. K., et. al., "Ultraviolet Disinfection of Wastewaters
  from Secondary Effluent and Combined Sewer Overflows."
  EPA-600/2-86-005, U.S. Environmental Protection Agency, WERL,
  Cincinnati, OH, 1986.
3. White, S. C., et. al., "A Study of Operational Ultraviolet
  Disinfection Equipment at Secondary Treatment Plants." Journal
  of the Water Pollution Control Federation, March 1986, Vol. 58,
  pp. 181-192.
Prepared by Environmental Resources Management, Inc.
For additional information contact:
EPA-OMPC (WH-595)               EPA-WERL (489)
401 M Street, SW                  26 West St. Clair Street
Washington, DC 20460               Cincinnati, OH 45268
(202) 382-7368/7369                 (513)  569-7688
 EPA Region 1
 John F. Kennedy Federal BuMng
 Boston, MA 02203

 EPA Region 2
 26 Federal Plaza
 New York, NY 10278

 EPA Region 3
 841 Chestnut Street
 Philadelphia, PA 19107

 EPA Region 4
 345 Courtland Street, NE
 Atlanta, GA 30365

 EPA Region 5
 230 South Dearborn Street
 Chicago, IL 60604
EPA Region 6
1201 Elm Street
Dallas, TX 75270

EPA Region 7
726 Minnesota Avenue
Kansas City, KS 66101

EPA Region 8
999 18th Street
Denver, CO 80202

EPA Region 9
215 Fremont Street
San Francisco, CA 94105

EPA Region 10
1200 6th Avenue
Seattle, WA 98101

-------