United States        Office of        Publication 9240.1 -06
         Environmental Protection    Solid Waste and     EPA-540/R-94/073
         Agency          Emergency Response    PB95-963503
                              August 1994

         Superfund
&EPA    USEPA CONTRACT
         LABORATORY PROGRAM

         STATEMENT OF WORK
         FOR ORGANIC ANALYSIS

         OLM03.1

-------
          Attachment A





USEPA CONTRACT LABORATORY PROGRAM







        STATEMENT OF WORK



               FOR



        ORGANICS ANALYSIS





Multi-Media, Multi-Concentration



             OLM03.0
        REVISION OLM03.1

           AUGUST 1994
       Chicago,
                      Protection



                      4.
                  DU
                                               PB95-963503

                                               EPA540/R-94/0

-------
                                          STATEMENT OF WORK

                                          TABLE OF CONTENTS



           EXHIBIT A:  SUMMARY OF REQUIREMENTS

           EXHIBIT B:  REPORTING AND DELIVERABLES  REQUIREMENTS

           EXHIBIT C:  TARGET COMPOUND  LIST  (TCL)  AND  CONTRACT  REQUIRED  QUANTITATION
                       LIMITS (CRQL)

           EXHIBIT D:  ANALYTICAL METHODS

S.         EXHIBIT E:  QUALITY ASSURANCE/QUALITY CONTROL  REQUIREMENTS
r v
           EXHIBIT F:  CHAIN-OF-CUSTODY, DOCUMENT  CONTROL AND STANDARD OPERATING
, '                      PROCEDURES

           EXHIBIT G:  GLOSSARY OF TERMS

;<'          EXHIBIT H:  DATA DICTIONARY  AND FORMAT  FOR  DATA DELIVERABLES  IN
-A                      COMPUTER-READABLE FORMAT
                                                                                   OLM03.0

-------
                              CONTRACTOR-OPERATED
                           SAMPLE MANAGEMENT OFFICE

The Sample Management Office  (SMO) is operated under a contract awarded and
administered by the U.S. Environmental Protection Agency (EPA).  Laboratory
contractors are advised that wherever in this document the words "Sample
Management Office" or "SMO" appear, EPA is referring to those contractor
employees.  The contract is currently held by DynCorp Viar, Inc. under
Contract No. 68-D9-0135.  Laboratory contractors are also advised that DynCorp
Viar employees are not representatives or agents of EPA.  As such, neither
DynCorp Viar nor its employees, nor any successor contractor, may change,
waive, or interpret any terms and conditions in this contract, including this
document OLM03.0.  All such questions or inquiries should be addressed to the
responsible party within EPA.
                                                                       OLM03.0

-------
       EXHIBIT A
SUMMARY OF REQUIREMENTS
          A-l                               OLM03.0

-------
                      Exhibit  A -  Summary  of  Requirements

                               Table  of  Contents


Section                                                                   Page

1.0   PURPOSE	    3

2.0   DESCRIPTION OF SERVICE   	    3

3.0   DATA USES	    3

4.0   SUMMARY OF REQUIREMENTS  	    3
      4.1   Introduction to the SOW	    3
      4.2   Overview of Major Task Areas	    4
      4.3   Technical and Management Capability   	    9
                                     A-2                               OLM03.0

-------
1. 0   PURPOSE

      The purpose of the multi-media, multi-concentration organic analytical
      service is to provide analytical data for use by the U.S. Environmental
      Protection Agency (EPA) in support of its investigation and clean-up
      activities under the Comprehensive Environmental Response, Compensation,
      and Liability Act of 1980 (CERCLA) and the Superfund Amendments and
      Reauthorization Act of 1986 (SARA).  Other EPA Program Offices who have
      similar analytical data needs also use this service.

2.0   DESCRIPTION OF SERVICE

      The organic analytical service provides a contractual framework for
      laboratories to apply EPA Contract Laboratory Program (CLP) analytical
      methods for the isolation, detection and quantitative measurement of 33
      volatile,  64 semivolatile, and 28 pesticide/Aroclor target compounds in
      water and soil/sediment environmental samples.  The analytical service
      provides the methods to be used, and the specific contractual
      requirements by which EPA will evaluate the data.  This service uses gas
      chromatography/mass spectrometry (GC/MS) and gas chromatography/electron
      capture (GC/EC) methods to analyze the target compounds.

3.0   DATA USES

      This analytical service provides data which EPA uses for a variety of
      purposes,  such as determining the nature and extent of contamination at
      a hazardous waste site, assessing priorities for response based on risks
      to human health and the environment, determining appropriate cleanup
      actions, and determining when remedial actions are complete.  The data
      may be used in all stages in the investigation of a hazardous waste site
      including site inspections,  Hazard Ranking System scoring, remedial
      investigations/feasibility studies, remedial design, treatability
      studies, and removal actions.   In addition, this service provides data
      that are available for use in Superfund enforcement/litigation
      activities.

4.0   SUMMARY OF REQUIREMENTS

4.1   Introduction to the SOW.  This statement of work (SOW) is designed as
      part of the documentation for a contract between EPA and a commercial
      laboratory performing analyses in support of EPA Superfund programs.
      The SOW comprises eight exhibits.  Exhibit A provides an overview of the
      SOW and its general requirements.  Exhibit B contains a description of
      the reporting and deliverables requirements, in addition to the data
      reporting forms and the forms instructions.  Exhibit C specifies the
      target compound list for this SOW with the contract-required
      quantitation limits for sample matrices.  Exhibit D details the specific
      analytical procedures to be used with this SOW and resulting contracts.
      Exhibit E provides descriptions of required quality assurance/quality
      control (QA/QC),  standard operating procedures, and procedures used for
      evaluating analytical methodologies, QA/QC performance, and the
      reporting of data.  Exhibit "F contains chain-of-custody and sample
      documentation requirements which the Contractor shall follow.  To ensure
                                      A-3                               OLM03.0

-------
      proper understanding of the terms utilized in this SOW,  a glossary can
      be found in Exhibit G.  (When a term is used in the text without
      explanation, the glossary meaning shall be applicable.)   Specifications
      for reporting data in computer-readable form appear in Exhibit H.

4.2   Overview of Major Task Areas.  For each sample, the Contractor shall
      perform the tasks described in this section.  Specific requirements for
      each task are detailed in the exhibits as referenced.

4.2.1    Task I:  Chain-of-Custody

4.2.1.1     Chain-of-Custody.  The Contractor shall receive and maintain
            samples under proper chain-of-custody procedures.   All associated
            document control and inventory procedures shall be developed and
            followed.  Documentation, as described therein,  shall be required
            to show that all procedures are being strictly followed.  This
            documentation shall be reported as the Complete Sample Delivery
            Group File (CSF) (see Exhibit B).  The Contractor shall establish
            and use appropriate procedures to handle confidential information
            received from the Agency.  See Exhibit F for specific
            requirements.

4.2.1.2     Sample Scheduling/Shipments.  Sample shipments to the Contractor's
            facility will be scheduled and coordinated by the CLP Sample
            Management Office (SMO).  The Contractor shall communicate with
            SMO personnel by telephone, as necessary throughout the process of
            sample scheduling,  shipment, analysis and data reporting, to
            ensure that samples are properly processed.

4.2.1.2.1      Samples will be shipped routinely to the Contractor through an
               overnight delivery service.  However, as necessary, the
               Contractor shall be responsible for any handling or processing
               required for the receipt of sample shipments, including pick-up
               of samples at the nearest servicing airport,  bus station or
               other carrier service within the Contractor's geographical
               area.  The Contractor shall be available to receive sample
               shipments at any time the delivery service is operating,
               including Saturdays.

4.2.1.2.2      If there are problems with the samples  (e.g., mixed media,
               containers broken or leaking) or sample documentation/paperwork
               (e.g., Traffic Reports not with shipment, sample and Traffic
               Report numbers do not correspond), the Contractor shall
               immediately contact SMO for resolution.  The Contractor shall
               immediately notify SMO regarding any problems and laboratory
               conditions that affect the timeliness of analyses and data
               reporting.  In particular, the Contractor shall notify  SMO
               personnel in advance regarding sample data that will be
               delivered late and shall specify the estimated delivery date.

4.2.1.2.3      To more effectively monitor the temperature of the sample
               shipping cooler, each USEPA Regional office may include a
               sample shipping cooler temperature blank with each cooler
               shipped.  The temperature blank will be clearly labeled:  USEPA
               COOLER TEMPERATURE  INDICATOR.

                                      A-4                              OLM03.0

-------
4.2.1.2.3.1    When the USEPA Regional office supplies a cooler temperature
               indicator bottle in the sample shipping cooler, the Contractor
               shall use the USEPA supplied cooler temperature indicator
               bottle to determine the cooler temperature.  The temperature of
               the cooler shall be measured at the time of sample receipt by
               the Contractor.

4.2.1.2.3.2    The temperature of the sample shipping cooler shall be measured
               and recorded immediately upon opening the cooler, and prior to
               unpacking the samples or removing the packing material.

4.2.1.2.3.3    To determine the temperature of the cooler, the contractor
               shall locate the cooler temperature indicator bottle in the
               sample shipping cooler, remove the cap and insert a calibrated
               thermometer into the cooler temperature indicator bottle.
               Prior to recording the temperature, the Contractor shall allow
               a minimum of 3 minutes, but not greater than 5 minutes for the
               thermometer to equilibrate with the liquid in the bottle.  At a
               minimum, the calibrated thermometer (±.1^_C) shall have a
               measurable range of 0 to 50 degrees Celsius.  Other devices
               which can measure temperature may be used if they can be
               calibrated to ±_1^_C and have a range of 0 to 20^_C.

4.2.1.2.3.4    If the temperature of the sample shipping cooler's temperature
               indicator exceeds 10 degrees Celsius,  the contractor shall
               contact the Sample Management Office (SMO) and inform them of
               the temperature deviation.  The SMO will contact the Region
               from which the samples were shipped for instructions on how to
               proceed.  The Region will either require that no sample
               analysis(es) be performed or that the Contractor proceed with
               the analysis(es).  The SMO will in turn notify the Contractor
               of the Region's decision.  The Contractor shall document the
               Region's decision in the SDG narrative.  Also in the SDG
               narrative, the Contractor shall list by fraction, the USEPA
               sample number, all samples which were shipped in a cooler which
               exceeded 10 degrees Celsius.

4.2.1.2.3.5    The Contractor shall record the temperature of the cooler on
               the DC-1 Form, under Remark #8 - Sample Conditions, and in the
               SDG narrative.

4.2.1.2.4      The Contractor shall accept all samples scheduled by SMO,
               provided that the total number of samples received in any
               calendar month does not exceed the monthly limitation expressed
               in the contract.  Should the Contractor elect to accept
               additional samples, the Contractor shall remain bound by all
               contract requirements for analysis of those samples accepted.

4.2.1.2.4      The Contractor shall be required to routinely return sample
               shipping containers (e.g., coolers) to the appropriate sampling
               office within 14 calendar days following shipment receipt (see
               Clause entitled Government Furnished Supplies and Materials).

4.2.2    Task II:  Analysis of Samples

4.2.2.1     Overview.  Sample analyses will be scheduled by groups of samples,
            each defined as a Case and identified by a unique EPA Case number
                                      A-5                               OLM03.1

-------
            assigned by SMO.  A Case signifies a group of samples collected at
            one site or geographical area over a finite time period, and will
            include one or more field samples with associated blanks.  Samples
            may be shipped to the Contractor in a single shipment or multiple
            shipments over a period of time, depending on the size of the
            Case.

4.2.2.1.1      A Case consists of one or more Sample Delivery Group(s).  A
               Sample Delivery Group (SDG) is defined by the following,
               whichever is most frequent:

               •   Each Case of field samples received, OR

               •   Each 20 field samples within a Case, OR

               •   Each 14 calendar day period (7 calendar day period for 14-
                   day data turnaround contracts) during which field samples
                   in a Case are received (said period beginning with the
                   receipt of the first sample in the SDG).

4.2.2.1.2      Samples may be assigned to SDGs by matrix (i.e.,  all soils in
               one SDG, all waters in another), at the discretion of the
               laboratory.  Such assignment shall be made at the time the
               samples are received,  and shall not be made retroactively.

4.2.2.2     Preparation Techniques.  Prepare samples as described in Exhibit
            D.   For semivolatile and pesticide/Aroclor samples,  an aliquot is
            ext.  .cted with a solvent and concent ated.  The concentrated
            extract is subjected to fraction-specific cleanup procedures and
            then analyzed by GC/MS for semivolatile or GC/EC for the
            pesticide/Aroclor target compounds listed in Exhibit C.  For
            volatile samples, an aliquot is purged with an inert gas, trapped
            on a solid sorbent, and then desorbed onto the GC/MS for analysis
            of the target compounds listed in Exhibit C.

4.2.2.3     Analytical Techniques.  The target compounds listed in Exhibit C
            shall be identified as described in the methodologies given in
            Exhibit D.  Automated computer programs may be used to facilitate
            the identification of compounds.

4.2.2.4     Qualitative Verification of Compounds.  The volatile and
            semivolatile compounds identified by GC/MS techniques shall be
            verified by an analyst competent in the interpretation of mass
            spectra by comparison of the suspect mass spectrum to the mass
            spectrum of a standard of the suspected compound.  This procedure
            requires the use of multiple internal standards.

4.2.2.4.1      If a compound initially identified by GC/MS techniques cannot
               be verified, but in the technical judgment of the mass spectral
               interpretation specialist the identification is correct, then
               the Contractor shall report that identification and proceed
               with quantitation.

4.2.2.4.2      The pesticide/Aroclor compounds identified by GC/EC techniques
               shall be verified by an analyst competent in the interpretation

                                     A-6                               OLM03.0

-------
               of gas chromatograms and by comparison of the retention times
               of the suspected unknowns with the retention times of
               respective standards of the suspected compounds.   Compounds
               shall also be confirmed by GC/MS techniques if the compounds
               are of sufficient concentration to be detected by the GC/MS.

4.2.2.5     Quantitation of Verified Compounds.  The Contractor shall
            quantitate components identified by GC/MS techniques by the
            internal standard method stipulated in Exhibit D.  Where multiple
            internal standards are required by EPA, the Contractor shall
            perform quantitation utilizing the internal standards specified in
            Exhibit D.  The Contractor shall quantitate components analyzed by
            GC/EC techniques by the external standard method stipulated in
            Exhibit D.  The Contractor shall also perform an initial three-
            point calibration, verify its linearity, determine the breakdown
            of labile components, and determine calibration factors for all
            standards analyzed by GC/EC techniques as described in Exhibit D.

4.2.2.6     Tentative Identification of Non-Target Sample Components.  For
            each analysis of a sample, the Contractor shall conduct mass
            spectral library searches to determine tentative compound
            identifications as follows.  For each volatile sample, the
            Contractor shall conduct a search to determine the possible
            identity of up to 30 organic compounds of greatest concentration
            which are not system monitoring compounds or internal standards
            and are not listed in Exhibit C under volatiles or semivolatiles.
            For each semivolatile sample, the Contractor shall conduct a
            search to determine the possible identification of up to 30
            organic compounds of greatest concentration which are not
            surrogates or internal standards and are not listed in Exhibit C
            under volatiles or semivolatiles.  In performing searches, the
            NIST/EPA/NIH (May 1992 release or later) and/or Wiley (1991
            release or later), or equivalent, mass spectral library shall be
            used.

            NOTE: Substances with responses less than 10% of the nearest
            internal standard are not required to be searched in this fashion.

4.2.2.7     Quality Assurance/Quality Control Procedures.  The Contractor
            shall strictly adhere to all specific QA/QC procedures prescribed
            in Exhibits D and E.  Records documenting the use of the protocol
            shall be maintained in accordance with the document control
            procedures prescribed in Exhibit F, and shall be reported in
            accordance with Exhibit B and Exhibit H.

4.2.2.7.1      The Contractor shall maintain a Quality Assurance Plan (QAP)
               with the objective of providing sound analytical chemical
               measurements.  This program shall incorporate the quality
               control procedures, any necessary corrective action,  and all
               documentation required during data collection as well as the
               quality assessment measures performed by management to ensure
               acceptable data production.

4.2.2.7.2      Additional quality control shall be conducted in the form of
               the analysis of laboratory evaluation samples submitted to the

                                      A-7                               OLM03.0

-------
               laboratory by the Agency.   The results of all such quality
               control or laboratory evaluation samples may be used as the
               basis for an equitable adjustment to reflect the reduced value
               of the data to the Agency  or rejection of data for:   sample(s),
               a fraction within an SDG,  or the entire SDG, and/or may be used
               as the basis for contract  action.  "Compliant performance" is
               defined as that which yields correct compound identification
               and concentration values as determined by the Agency,  as well
               as meeting the contract requirements for analysis (Exhibit D),
               quality assurance/quality  control (Exhibit E), data reporting
               and other deliverables (Exhibits B and H), and sample custody,
               sample documentation and standard operating procedure
               documentation (Exhibit F).

4.2.3    Task III:  Reporting Requirements

4.2.3.1     EPA has provided to the Contractor formats for the reporting of
            data (Exhibits B and H).  The Contractor shall be responsible for
            completing and returning analysis data sheets and submitting
            computer-readable data on diskette in the format specified in this
            SOW and within the time specified in the Contract Performance/
            Delivery Schedule in Exhibit  B.

4.2.3.2     Use of formats other than those designated by EPA will be deemed
            as noncompliant.   Such data are unacceptable.  Resubmission in the
            specified format at no additional cost to the Agency shall be
            required.

4.2.3.3     Computer-generated forms may  be submitted in the hardcopy data
            package(s) provided that the  forms are in exact EPA format.  This
            means that the order of data  elements is the same as on each EPA-
            required form, including form numbers and titles, page numbers and
            header information.

4.2.3.4     The data reported by the Contractor on the hardcopy data forms and
            the associated computer-readable data submitted by the Contractor
            shall contain identical information.  If discrepancies are found
            during government inspection, the Contractor shall be required to
            resubmit either the hardcopy forms or the computer-readable data,
            or both sets of data, at no additional cost to the Agency.
                                      A-8                              OLM03.0

-------
4.3   Technical and Management Capability

4.3.1    Personnel.  The Contractor shall have adequate personnel at all times
         during the performance of the contract to ensure that EPA receives
         data that meet the terms and conditions of the contract.

4.3.2    Instrumentation.  The Contractor shall have sufficient gas
         chromatograph/electron capture/data system (GC/EC/DS), gas
         chromatograph/mass spectrometer/data system (GC/MS/DS), including
         magnetic tape storage devices, and gel permeation chromatography
         system (GPC) capability to meet all the terms and conditions of the
         contract.

4.3.3    Facilities.  The Contractor shall maintain a facility suitable for
         the receipt, storage, analysis, and delivery of the product meeting
         the terms and conditions of the contract.
                                     A-9                               OLM03.0

-------
               EXHIBIT B
REPORTING AND DELIVERABLES REQUIREMENTS
                  B-l                               OLM03.0

-------
             Exhibit  B  -  Reporting and Deliverables Requirements

                               Table of Contents
Section                                                                   Page

1.0   CONTRACT REPORTS/DELIVERABLES DISTRIBUTION  	    3
      1.1   35-Day Data Turnaround Contracts  	    3
      1.2   14-Day Data Turnaround Contracts  	    6
      1.3   Distribution  	    9

2.0   REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES 	   10
      2.1   Introduction	   10
      2.2   Resubmission of Data  .   	   10
      2.3   Quality Assurance Plan and Standard Operating Procedures  .  .   11
      2.4   Sample Traffic Reports  	   11
      2.5   Sample Data Summary Package 	   12
      2.6   Sample Data Package	   12
      2.7   Complete SDG File	   28
      2.8   Data in Computer-Readable Form	   29
      2.9   GC/MS Tapes	   30
      2.10  Extracts	   30

3.0   FORMS INSTRUCTIONS  	   31
      3.1   Introduction	   31
      3.2   General Information 	   31
      3.3   Header Information  	   32
      3.4   Organic Analysis Data Sheet (Form I, All Fractions) 	   36
      3.5   Organic Analysis Data Sheet:   Tentatively Identified
            Compounds (Form I VOA-TIC and Form I SV-TIC)  	   41
      3.6   System Monitoring Compound Recovery (Form II VOA)  	   42
      3.7   Surrogate Recovery (Form II SV and Form II PEST)	   43
      3.8   Matrix Spike/Matrix Spike Duplicate Recovery (Form III, All
            Fractions)  	   45
      3.9   Method Blank Summary (Form IV, All Fractions) 	   46
      3.10  GC/MS Instrument Performance Check and Mass Calibration
            (Form V VOA and Form V SV)  	   47
      3.11  GC/MS Initial Calibration Data (Form VI VOA and Form VI SV)  .   49
      3.12  GC/EC Initial Calibration Data (Form VI PEST) 	   50
      3.13  GC/MS Continuing Calibration Data (Form VII VOA and Form VII
            SV)	   52
      3.14  GC/EC Calibration Verification Summary  (Form VII,  PEST-1,
            PEST-2)	   53
      3.15  Internal Standard Area and RT Summary (Form VIII VOA and
            Form VIII SV)	   54
      3.16  Pesticide Analytical Sequence (Form VIII PEST)   	   56
      3.17  Pesticide Cleanup Summary (Form IX, PEST-1, PEST-2) 	   58
      3.18  Pesticide/Aroclor Identification  (Form X, PEST-1,  PEST-2) .  .   59
      3.19  Sample Log-In Sheet (Form DC-1) 	   61
      3.20  Document Inventory Sheet  (Form DC-2)  	   62

4.0   DATA REPORTING FORMS	   63
                                      B-2                              OLM03.0

-------
                                                           Exhibit B—Section 1
                                     Contract Reports/Deliverables Distribution
 1.0   CONTRACT REPORTS/DELIVERABLES DISTRIBUTION

 1.1   35-Day Data Turnaround Contracts.   The following table reiterates the
       contract reporting and deliverables requirements specified in the
       Contract Schedule (Performance/Delivery Schedule)  and specifies the
       distribution that is required for  each deliverable.

       NOTE:  Specific recipient  names and addresses are subject to change
       during the term of the contract.   The Administrative Project officer
       (APO)  will notify the Contractor in writing of such  changes when they
       occur.

                                    Table 1
            Item
                No. of
                Copies
Delivery Schedule
                                                              Distribution

                                                                c     j
                                                                o     -~.
                                                            O   DI    w
                                                            SO)     S
                                                            OT   OS     W
 A.
 B.
 C.
 D.
 E.
Sample Traffic
Reports
Sample Data
Summary
Package

Sample Data
Package


Data in
Computer
Readable Form

Complete SDG
File
5 working days
after receipt of
last  sample in
Sample Delivery
Group (SDG).**

35 days after
receipt of last
sample in SDG.

35 days after
receipt of last
sample in SDG.

35 days after
receipt of last
sample in SDG.
35 days after
receipt of last
sample in SDG.
      Contractor-concurrent delivery to EMSL/LV may be  required upon request by
the APO and/or EMSL/LV.  Retain for 365 days after  data submission, and submit
as directed within  7  days after receipt of written request by the APO and/or
EMSL/LV.
                                      B-3
                                                                       OLM03.0

-------
Exhibit B—Section 1
Contract Reports/Deliverables Distribution
            Item
                                                            Distribution
                 No. of   Delivery
                 Copies   Schedule
                                                          §
                                                          w
                                                     c
                                                     O
                                                     o<
                                                     &
                                      OT
                                      a
                                      H
F. Standard
Operating
Procedures —
Technical and
Evidentiary
1 Submit within 7
days of receipt
of written
request to
recipients as
directed.
 G.
Quality
Assurance Plan
       Submit within 7
       days of receipt
       of written
       request ti
       recipients as
       directed.
 H.
GC/MS Tapes
Lot    Retain for 365
       days after data
       submission.
As directed
 I.
           Extracts
                              Lot
                         Submit within  7
                         days  after
                         receipt of
                         written request
                         by APO and/or
                         EMSL/LV.

                         Retain for  365
                         days  after  data
                         submission.
                               As  directed
                                     Submit within 7
                                     days after
                                     receipt of
                                     written request
                                     by APO or SMO,
                                     at the Agency's
                                     direction.
                                     B-4
                                                            OLM03.0

-------
                                                          Exhibit B—Section 1
                                    Contract Reports/Deliverables Distribution
Footnotes;

     Also required in the Sample Data Summary Package.

     A sample delivery group  (SDG) is a group of samples within a Case,
     received over a period of 14 days or less (7 days or less for 14-day data
     turnaround contracts) and not exceeding 20 samples.  Data for all samples
     in the SDG are due concurrently.  The date of delivery of the SDG or any
     samples within the SDG is the date that the last sample in the SDG is
     received.  (See Exhibit A for further description.)

***   DELIVERABLES ARE TO BE REPORTED TOTAL AND COMPLETE.  Concurrent delivery
     required.  Delivery shall be made such that all designated recipients
     receive the item on the same calendar day.  This includes resubmission of
     both the hardcopy and diskette.  The date of delivery of the SDG, or any
     sample within the SDG, is the date all samples have been delivered.  If
     the deliverables are due on a Saturday, Sunday or Federal holiday, then
     they shall be delivered on the next business day.  Deliverables delivered
     after this time will be considered late.

     Complete SDG File will contain the original sample data package plus all
     of the original documents described under Section 2.7.

     See Exhibit E and Exhibit F for a more detailed description.

NOTE: As specified in the Contract Schedule (Government Burnished Supplies and
Materials),  unless otherwise instructed by the CL  Sample Management Office
based on a Regional decision,  the Contractor shall dispose of unused sample
volume and used sample bottles/containers no earlier than sixty (60) days
following submission of the reconciled Complete SDG File.  Sample disposal and
disposal of unused sample bottles/containers is the responsibility of the
Contractor and should be done in accordance with all applicable laws and
regulations governing disposal of such materials.
                                      B-5                               OLM03.0

-------
Exhibit B—Section  1
Contract Reports/Deliverables Distribution
 1.2   14-Day Data Turnaround Contracts.  The following table reiterates  the
      contract reporting and deliverables requirements specified  in the
      Contract Schedule  (Performance/Delivery Schedule) and specifies  the
      distribution that is required for each deliverable.

      NOTE: Specific recipient names and addresses are subject to  change
      during the term of the contract.  The Administrative Project Officer
      (APO) will notify the Contractor in writing of such changes  when they
      occur.

                                    Table 2
            Item
                No. of
                Copies
Delivery
Schedule
                                                             Distribution
                                                          O
                                                          w
c
o
-H
CP
0)
ct
                                                                      ,J
w
x
W
 A.
 B.
 C.
 D.
 E.
Sample Traffic
Reports
Sample Data
Summary
Package

Sample Data
Package

Data in
Computer
Readable Form
Complete SDG
File
5 working days
after receipt of
last sample in
Sample Delivery
Group (SDG).**
14 days after
receipt of last
sample in SDG.

14 days after
receipt of last
sample in SDG.
14 days after
receipt of last
sample in SDG.
14 days after
receipt of last
sample in SDG.
                                                               X
     Contractor-concurrent delivery to EMSL/LV may be required upon request by
the APO and/or EMSL/LV.  Retain for 365 days after data submission, and submit
as directed within  7  days after receipt of written  request  by the APO and/or
EMSL/LV.
                                      B-6
                                                          OLM03.0

-------
                                                          Exhibit B—Section 1
                                    Contract  Reports/Deliverables Distribution
                                                            Distribution
Item
F. Standard
Operating
Procedures —
Technical and
Evidentiary
No. of -H
O O1
Copies Delivery Schedule S 
-------
Exhibit B—Section 1
Contract Reports/Deliverables Distribution
Footnotes;

     Also required in the Sample Data Summary Package.

     A sample delivery group (SDG) is a group of samples within a Case,
     received over a period of 14 days or less (7 days or less for 14-day data
     turnaround contracts) and not exceeding 20 samples.  Data for all samples
     in the SDG are due concurrently.  The date of delivery of the SDG or any
     samples within the SDG is the date that the last sample in the SDG is
     received.  (See Exhibit A for further description.)

***   DELIVERABLES ARE TO BE REPORTED TOTAL AND COMPLETE.  Concurrent delivery
     required.  Delivery shall be made such that all designated recipients
     receive the item on the same calendar day.  This includes resubmission of
     both the hardcopy and diskette.  The date of delivery of the SDG, or any
     sample within the SDG, is the date all samples have been delivered.  If
     the deliverables are due on a Saturday, Sunday or Federal holiday, then
     they shall be delivered on the next business day.  Deliverables delivered
     after this time will be considered late.

     Complete SDG File will contain the original sample data package plus all
     of the original documents described under Section 2.7.

     See Exhibit E and Exhibit F for a more detailed description.

NOTE: As specified in the Contract Schedule (Government Furnished Supplies and
Materials),  unless otherwise instructed by the CLP Sample Management Office
based on a Regional decision,  the Contractor shall dispose of unused sample
volume and used sample bottles/containers no earlier than sixty (60) days
following submission of the reconciled Complete SDG File.  Sample disposal and
disposal of unused sample bottles/containers is the responsibility of the
Contractor and should be done in accordance with all applicable laws and
regulations governing disposal of such materials.
                                      B-8                               OLM03.0

-------
                                                          Exhibit B—Section 1
                                    Contract Reports/Deliverables Distribution
1.3   Distribution.  The following addresses correspond to the "Distribution"
      column in the tables in Sections 1.1 and 1.2.
     SMO:
     Region:
     EMSL/LV:
USEPA Contract Laboratory Program
Sample Management Office (SMO)
P. O. Box 818
Alexandria, VA  22313

For overnight delivery service, use street address:
300 North Lee Street, Suite 200
Alexandria, VA  22314

USEPA Region:  The Sample Management Office will provide the
Contractor with the list of addresses for the 10 EPA Regions.
SMO will provide the Contractor with updated Regional
address/name lists as necessary throughout the period of the
contract and identify other client recipients on a
case-by-case basis.

USEPA Environmental Monitoring Systems Laboratory (EMSL/LV)
ATTN:  Data Audit Staff
P. 0. Box 93478
Las Vegas, NV  89193-3478

For overnight delivery service, use street address:
944 E.  Harmon,  Executive Center
Las Vegas, NV  89109
                                     B-9
                                                                       OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
2.0   REPORTING REQUIREMENTS AND ORDER OF DATA DELIVERABLES

2.1   Introduction.  The Contractor shall provide reports and other
      deliverables as specified in the Contract Schedule (Performance/Delivery
      Schedule).  The required content and form of each deliverable is
      described in this exhibit.  All reports and documentation must be:

      •   Legible,

      •   Clearly labeled and completed in accordance with instructions in
          this exhibit,

      •   Arranged in the order specified in this section,

      •   Paginated consecutively in ascending order starting from the SDG
          Narrative,  and

      •   Copies must be legible and double-sided.

      NOTE:  Complete SDG files need not be double-sided.  (The CSF is composed
      of original documents.)  However, sample data packages delivered to SMO
      and upon request to EMSL/LV must be double-sided.

2.1.1    Requirements for each deliverable item cited in the Contract Schedule
         (Contract Performance/Delivery Schedule) are specified in Sections
         2.3-2.10.  Prior to submission,  the Contractor =;hall arrange items
         and the components of each item in the 03der listed in these
         sections.

2.1.2    The Contractor shall use EPA Case numbers (including SDG numbers) and
         EPA sample numbers to identify samples received under this contract,
         both verbally and in reports/correspondence.  The contract number
         shall be specified in all correspondence.

2.2   Resubmission of Data.  If submitted documentation does not conform to
      the above criteria, the Contractor shall resubmit such documentation
      with deficiency(ies) corrected, at no additional cost to the Agency.

2.2.1    The Contractor shall respond within seven (7) days to written
         requests from data recipients for additional information or
         explanations that result from the Government's inspection activities
         unless otherwise specified in the contract.

2.2.2    Whenever the Contractor is required to submit or resubmit data as a
         result of an on-site laboratory evaluation, or through an
         Administrative Project Officer/Technical Project Officer action, or
         through a Regional data reviewer's request, the data shall be clearly
         marked as ADDITIONAL DATA and shall be sent to both contractual data
         recipients (SMO and the Region; to EMSL/LV upon written request).
         The Contractor shall include a cover letter which describes which
         data are being delivered, to which EPA Case(s) the data pertain, and
         who requested the data.
                                     B-10                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables


2.2.3    Whenever the Contractor is required to submit or resubmit data as a
         result of Contract Compliance Screening (CCS) review by SMO, the data
         shall be sent to both contractual data recipients (SMO and the
         Region; to EMSL/LV when a written request for the sample data package
         has been made).  In all instances, the Contractor shall include a
         color-coded COVER SHEET (Laboratory Response To Results of Contract
         Compliance Screening) provided by SMO.

2.3   Quality Assurance Plan and Standard Operating Procedures.  The
      Contractor shall adhere to the requirements in Exhibits E and F.

2.4   Sample Traffic Reports.  Each sample received by the Contractor will be
      labeled with an EPA sample number, and will be accompanied by a Sample
      Traffic Report (TR) bearing the sample number and descriptive
      information regarding the sample.  The Contractor shall complete the TR
      (marked "Lab Copy for Return to SMO"), recording the date of sample
      receipt and sample condition upon receipt for each container, and shall
      sign the TR.  Information shall be recorded for each sample in the SDG.

2.4.1    The Contractor shall submit TRs in SDG sets (i.e., TRs for all
         samples in an SDG shall be clipped together), with an SDG cover sheet
         attached.  The SDG cover sheet shall contain the following items:

         •   Laboratory name,

         •   Contract number,

         •   Sample analysis price (full sample price from the contract),

         •   Case number, and

         •   List of EPA sample numbers of all samples in the SDG, identifying
             the first and last samples received, and their dates of receipt
             (LRDs).

             NOTE:  When more than one sample is received in the first or last
             SDG shipment, the "first" sample received would be the lowest
             sample number (considering both alpha and numeric designations);
             the "last" sample received would be the highest sample number
             (considering both alpha and numeric designations).

2.4.2    Each TR shall be clearly marked with the SDG number, entered below
         the laboratory receipt date on the TR.  The TR for the last sample
         received in the SDG shall be clearly marked "SDG—FINAL SAMPLE."  The
         SDG number is the EPA sample number of the first sample received in
         the SDG.  When several samples are received together in the first SDG
         shipment, the SDG number shall be the lowest sample number
         (considering both alpha and numeric designations) in the first group
         of samples received under the SDG.
                                     B-ll                              OLM03.0

-------
 Exhibit B—Section 2
 Reporting Requirements  and  Order  of  Data Deliverables
 2.4.3     If  samples  are  received  at the  laboratory with multi-sample  TRs,  all
          the samples on  one multi-sample TR may not necessarily be  in the  same
          SDG.   In  this instance,  the  Contractor shall make the appropriate
          number of photocopies  of the TR,  and submit one copy with  each  SDG
          cover  sheet.

 2.5    Sample Data  Summary  Package.  The  sample data summary package shall  be
       ordered as follows and  shall be submitted separately (i.e., separated by
       rubber bands,  clips  or  other means)  directly preceding the sample  data
       package.   Sample data forms shall  be arranged in increasing EPA sample
       number order,  considering both  letters and numbers.  For example,  BE400
       is  a lower sample  number  than BF100, as E precedes F in the alphabet.
       The SDG number shall be reported on  all data reporting forms.   The
       sample data  summary package shall  contain data for all samples  in  one
       SDG of the Case, as  follows.  (See Section 2.6 for a detailed
       description  of  each  item.)  The sample data summary package shall  be
       arranged  in  the same manner as  the sample data package.

 2.5.1     SDG Narrative.

 2.5.2     Arranged  by  fraction and by  sample within each fraction:   tabulated
          target  compound results  (Form I)  for the volatile, semivolatile and
          pesticide fractions  and  tentatively identified compounds (Form  I TIC)
          for  the volatile and semivolatile fractions only.

 2.5.3     Arranged  by  fraction:  system monitoring compound or surrogate  spike
          analys  3  results  (Form II)  by matrix (water and/or soil) and  for
          soil, by  concentration (low  or  medium),  for the volatile,
          semivolatile and pesticide fractions.

 2.5.4    Arranged  by fraction:  matrix spike/matrix spike duplicate results
          (Form III)  for the volatile,  semivolatile and pesticide fractions.

 2.5.5    Arranged by fraction:  blank data (Form IV)  and tabulated  results
          (Form I)  for the volatile,  semivolatile and pesticide fractions
          including tentatively identified compounds (Form I TIC)  for the
         volatile and semivolatile fractions only.

2.5.6    Arranged by fraction:  internal standard area data (Form VIII)  for
         the volatile and semivolatile fractions only.

2.6  Sample Data Package.  The sample data package is divided into the five
     major units described in this section.  The last three units are each
     specific to an analytical fraction  (yolatiles,  semivolatiles,  and
     pesticides/Aroclors).   If the analysis of  a fraction is  not required,
     then that fraction-specific unit is not required as a deliverable.  The
     sample data package  shall include data for the analyses  of all samples in
     one SDG, including field samples, dilutions,  reanalyses,  blanks,  matrix
     spikes,  and matrix spike duplicates.  The  Contractor shall retain a copy
     of the sample data package for  365 days after final acceptance of data.
     After this  time, the Contractor may dispose of the package.

2.6.1    SDG Narrative.   This document shall be clearly labeled "SDG
         Narrative" and shall contain:   laboratory name;  Case number;  EPA
         sample  numbers in the SDG,  differentiating between initial analyses
         and reanalyses;  SDG number;  Contract number;  and detailed
                                     B-12                              OLM03.1

-------
                                                           Exhibit  B—Section  2
                          Reporting  Requirements  and  Order  of  Data  Deliverables
                                                                     Volatiles
          documentation of any quality  control,  sample,  shipment  and/or
          analytical  problems  encountered  in  processing  the  samples  reported  in
          the  data  package.  All  GC columns used for analysis  shall  be
          documented  here, by  fraction.  List the GC column  identification—
          brand name,  the internal  diameter,  in  mm,  and  the  length,  in meters,
          packing/coating material  and  film thickness.   The  trap  used for
          volatile  analysis shall be described here.  List trap name, when
          denoted by  the manufacturer,  its composition  (packing material/brand
          name,  amount of packing material, in length, cm).  All  tentatively
          identifed alkanes and their estimated  concentrations are to be
          reported  here.   The  EPA sample number, the CAS number,  when
          available,  the alkane compound (or  series)  name, and its estimated
          concentration shall  be  provided  in  a tabular format.  The  Contractor
          shall  include any technical and  administrative problems encountered,
          the  corrective actions  taken, the resolution,  an explanation for all
          flagged edits (e.g.,  manual edits)  on  quantitation lists.   The
          Contractor  shall document in  the SDG Narrative all instances of
          manual  integration.   The  SDG  Narrative shall contain the following
          statement,  verbatim:  "I  certify that  this data package is in
          compliance  with the  terms and conditions of the contract,  both
          technically  and for  completeness, for  other than the conditions
          detailed  above.   Release  of the  data contained in  this  hardcopy data
          package and  in the computer-readable data  submitted  on  diskette has
          been  authorized by the  laboratory manager  or his designee,  as
          verified  by  the following signature."  This statement shall be
          directly  followed  by  an original signature of  the  laboratory manager
          or his  designee with  a  typed  line below it  containing the  signer's
          name  and  title,  and  the date  of  signature.

2.6.1.1     Whenever  data from sample  reanalyses are submitted,  the Contractor
            shall  state  in  the SDG Narrative  for each reanalysis whether the
            reanalysis is billable,  and if so, why.

2.6.1.2     The  Contractor  shall list  the pH  determined for each water sample
            submitted  for volatiles  analysis.  This  information  may appear as
            a simple  list or table in  the SDG Narrative.  The purpose of this
            pH determination  is  to ensure that all  water volatiles  samples
            were acidified  in  the  field.  No  pH adjustment  is to be performed
            by the Contractor  on water samples for  volatiles analysis.

2.6.2     Traffic Reports.   The Contractor shall include a copy of the TRs
          submitted in  Section  2.4  for  all of  the samples in the  SDG.  The TRs
          shall be  arranged  in  increasing  EPA  sample  number  order, considering
          both letters  and numbers.   Copies of the SDG cover sheet are to be
          included with the  copies  of the  TRs.   (See  Section 2.4  for more
          detail  on reporting requirements for TRs.)  In the case of multi-
          sample  TRs,  the  Contractor  shall make the  appropriate nuir.ber of
          photocopies of  the TR so  that a  copy is submitted  with  each
          applicable data  package.   In  addition,  in  any  instance where samples
          from more than  one multi-sample  TR are in  the  same data package, the
          Contractor shall submit a  copy of the SDG  cover sheet with copies of
          the TRs.

2.6.3    Volatiles Data

2.6.3.1     Volatiles  QC  Summary

2.6.3.1.1      System  Monitoring Compound Summary (Form II  VGA).

2.6.3.1.2      Matrix  Spike/Matrix Spike  Duplicate Summary  (Form III VOA) .
                                     B-13                              OLM03.1

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
Volatiles
2.6.3.1.3      Method Blank Summary  (Form IV VOA):  If more than a single form
               is necessary, forms shall be arranged in chronological order by
               date of analysis of the blank, by instrument.

2.6.3.1.4      GC/MS instrument performance check (Form V VOA):  If more than
               a single form is necessary, forms shall be arranged in
               chronological order, by instrument.

2.6.3.1.5      Internal Standard Area and RT Summary (Form VIII VOA):  If more
               than a single form is necessary, forms shall be arranged in
               chronological order, by instrument.

2.6.3.2     Volatiles Sample Data.  Sample data shall be arranged in packets
            with the Organic Analysis Data Sheet (Form I VOA, including Form I
            VOA-TIC), followed by the raw data for volatile samples.  These
            sample packets shall be placed in increasing EPA sample number
            order,  considering both letters and numbers.

2.6.3.2.1      Target Compound Results, Organic Analysis Data Sheet (Form I
               VOA).  Tabulated results (identification and quantitation) of
               the specified target compounds (Exhibit C,  Volatiles) shall be
               included.  The validation and release of these results are
               authorized by a specific,  signed statement in the SDG Narrative
               (see Section 2.6.1).  In the event that the laboratory manager
               cannot verify all data reported for each sample, the laboratory
               manager shall provide a detailed description of the problems
               associated with the sample in the SDG Narrative.

2.6.3.2.2      Tentatively Identified Compounds (Form I VOA-TIC).  Form I
               VOA-TIC is the tabulated list of the highest probable match for
               up to 30 organic compounds that are not system monitoring
               compounds or internal standard compounds and are not listed in
               Exhibit C.   It includes the Chemical Abstracts Service (CAS)
               registry number (if applicable), tentative identification, and
               estimated concentration.  This form shall be included even if
               no compounds are found. If no compounds are found, indicate
               this on the form by entering "0" in the field for "Number
               Found."

2.6.3.2.3      Reconstructed Total Ion Chromatograms (for each sample or
               sample extract,  including dilutions and reanalyses) .
               Reconstructed ion chromatograms shall be normalized to the
               largest nonsolvent component and shall contain the following
               header information:

               •   EPA sample number,

               •   Date and time of analysis,

               •   GC/MS instrument identifier,
                                     B-14                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                                     Volatiles
                •   Lab  file  identifier, and

                •   Analyst ID.

2.6.3.2.3.1        Internal standards and system monitoring compounds  shall be
                   labeled with the names of compounds, either directly out
                   from  the peak or on a printout of retention times if
                   retention  times are printed over the peak.

2.6.3.2.3.2        If automated data system procedures are used for preliminary
                   identification and/or guantitation of the target compounds,
                   the complete data system report shall be included in all
                   sample data packages, in addition to the reconstructed  ion
                   chromatogram.  The complete data system report shall include
                   all of the information listed below.  For laboratories  which
                   do not use the automated data system procedures, a
                   laboratory "raw data sheet" containing the following
                   information shall be included in the sample data package, in
                   addition to the chromatogram:

                   •   EPA sample number,

                   •   Date and time of analysis,

                   •   Retention time or scan number of identified target
                      compounds,

                   •   Ion used for quantitation with measured area,

                   •   Copy of area table from data system,

                   •   GC/MS instrument identifier,

                   •   Lab file identifier,  and

                   •   Analyst ID.

2.6.3.2.3.3        In all instances where the data system report has been
                  edited, or where manual integration or quantitation has been
                  performed,  the GC/MS operator shall identify such edits or
                  manual procedures by initialing and dating the changes made
                  to the report, and shall include the integration scan range.
                   In addition,  a hardcopy printout of the EICP of the
                  quantitation ion displaying the manual integration shall be
                   included in the raw data.   This applies to all compounds
                   listed in Exhibit C (Volatiles),  internal standards and
                  system monitoring compounds.

                   •   EICPs displaying each manual integration.
                                     B-15                              OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
Volatiles
2.6.3.2.4      Other Required Information.   For each sample,  by each compound
               identified,  the following items shall be included in the data
               package.

               •   Copies of raw spectra and copies of background-subtracted
                   mass spectra of target compounds listed in Exhibit C
                   (Volatiles) that are identified in the sample and
                   corresponding background-subtracted target compound
                   standard mass spectra.  Spectra shall be labeled with EPA
                   sample number, lab file identifier, and date and time of
                   analysis, and GC/MS instrument identifier.  Compound names
                   shall be clearly marked on all spectra.

               •   Copies of mass spectra of organic compounds not listed in
                   Exhibit  C with associated best-match spectra (minimum of
                   one, maximum of three best matches).  Spectra shall be
                   labeled  with EPA sample number,  lab file identifier, and
                   date and time of analysis, and GC/MS instrument identifier.
                   Compound names shall be clearly marked on all spectra.

2.6.3.3     Volatiles Standards Data

2.6.3.3.1      Initial calibration data (Form VI VGA) shall be included in
               order by instrument, if more than one instrument is used.

               •   Volatile standard(s) reconstructed ion chromatograras and
                   guantitation reports for the initial (five-point)
                   calibration, labeled as in Section 2.6.3.2.3.  Spectra are
                   not required.

               •   All initial calibration data that pertain to samples in the
                   data package shall be included,  regardless of when it was
                   performed and for which Case.  When more than one initial
                   calibration is performed, the data shall be in
                   chronological o_der, by instrument.

               •   EICPs displaying each manual integration.

2.6.3.3.2      Continuing calibration data (Form VII VOA) shall be included in
               order by instrument, if more than one instrument is used.

               •   Volatile standard(s) reconstructed ion chromatograms and
                   quantitation reports for all continuing (12-hour)
                   calibrations, labeled as in Section 2.6.3.2.3.  Spectra are
                   not required.

               •   When more than one continuing calibration is performed,
                   forms shall be in chronological order, by instrument.

               •   EICPs displaying each manual integration.
                                     B-16                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                                     Volatiles
2.6.3.3.3      In all instances where the data system report has been edited,
               or where manual integration or guantitation has been performed,
               the GC/MS operator shall identify such edits or manual
               procedures by initialing and dating the changes made to the
               report, and shall include the integration scan range.  In
               addition, a hardcopy printout of the EICP of the guantitation
               ion displaying the manual integration shall be included in the
               raw data.  This applies to all compounds listed in Exhibit C
               (Volatiles), internal standards and system monitoring
               compounds.

2.6.3.4     Volatiles Raw QC Data

2.6.3.4.1      BFB data shall be arranged in chronological order by instrument
               for each 12-hour period, for each uC/MS system utilized.

               •   Bar graph spectrum, labeled as in Section 2.6.3.2.3.

               •   Mass listing, labeled as in Section 2.6.3.2.3.

               •   Reconstructed total ion chromatogram,  labeled as in Section
                   2.6.3.2.3.

2.6.3. 1.2      Blank data shall be arranged by type of blank (method, storage,
               instrument)  and shall be in chronological  order by instrument.

               NOTE:   This order is different from that used for samples.

               •   Tabulated results (Form I VOA).

               •   Tentatively identified compounds (Form I VOA-TIC) even if
                   none are found.

               •   Reconstructed ion chromatogram(s)  and  guantitation
                   report(s), labeled as in Section 2.6.3.2.3.

               •   Target compound spectra with laboratory-generated standard,
                   labeled as in Section 2.6.3.2.4.   Data systems which are
                   incapable of dual display shall  provide spectra in the
                   following order:

                       Raw target compound spectra.

                   —  Enhanced or background-subtracted  spectra.

                       Laboratory-generated standard spectra.
                                     B-17                              OLM03.0

-------
 Exhibit  B—Section  2
 Reporting  Requirements  and Order of Data Deliverables
 Semivolatiles
                •   GC/MS  library search spectra for tentatively identified
                   compounds, labeled as  in Section 2.6.3.2.4.

                •   Quantitation/calculation of tentatively identified  compound
                   concentrations.

2.6.3.4.3       Volatiles  Matrix Spike Data

                •   Tabulated results (Form I VOA) of target compounds. Form I
                   VOA-TIC is not required.

                •   Reconstructed ion chromatogram(s) and quantitation
                   report(s), labeled as  in Section 2.6.3.2.3.  Spectra are
                   not required.

2.6.3.4.4       Volatiles Matrix Spike Duplicate Data

                •   Tabulated results (Form I VOA) of target compounds. Form I
                   VOA-TIC is not required.

                •   Reconstructed ion chromatogram(s) and quantitation
                   report(s), labeled as  in Section 2.6.3.2.3.  Spectra are
                   not required.

2.6.4    Semivolatiles Data

2.6.4.1     Semivolatiles QC Summary

2.6.4.1.1       Surrogate Percent Recovery Summary (Form II SV).

2.6.4.1.2      Matrix Spike/Matrix Spike Duplicate Summary (Form III SV)

2.6.4.1.3      Method Blank Summary (Form IV SV):  If more than a single form
                is necessary,  forms shall be arranged in chronological order by
               date of analysis cf the blank,  by instrument.

2.6.4.1.4      GC/MS Instrument Performance Check (Form V SV):  If more than a
               single form is necessary, forms shall be arranged In
               chronological  order,  by instrument.

2.6.4.1.5       Internal Standard Area and RT Summary (Form VIII  SV):  If more
               than a single  form is necessary,  forms shall be arranged in
               chronological  order,  by instrument.

2.6.4.2     Semivolatiles Sample Data.   Sample data shall be arranged in
            packets with the  Organic Analysis  Data Sheet (Form I SV, including
            Form I SV-TIC),  followed by the raw data for semivolatile samples.
            These sample packets shall be placed in increasing EPA sample
            number order,  considering both letters and numbers.

2.6.4.2.1      Target Compound Results,  Organic Analysis Data Sheet (Form I
               SV-1,  SV-2).   Tabulated results (identification and
               quantitation)  of the specified  target compounds (Exhibit C,

                                     B-18                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                                 Semivolatiles
               Semivolatiles) shall be included.  The validation and release
               of these results are authorized by a specific, signed statement
               in the SDG Narrative (see Section 2.6.1).   In the event that
               the laboratory manager cannot verify all data reported for each
               sample, the laboratory manager shall provide a detailed
               description of the problems associated with the sample in the
               SDG Narrative.

2.6.4.2.2      Semivolatile Tentatively Identified Compounds (Form I SV-TIC).
               Form I SV-TIC is the tabulated list of the highest probable
               match for up to 30 of the non-surrogate/non-internal standard
               organic compounds that are not listed in Exhibit C (Volatiles,
               Semivolatiles).  It includes the CAS registry number (if
               applicable), tentative identification, and estimated
               concentration.  This form shall be included even if no
               compounds are found.  If no compounds are found, indicate this
               on the form by entering "0" in the field for "number found."

2.6.4.2.3      Reconstructed Total Ion Chromatograms (for each sample,
               including dilutions and reanalyses).  Reconstructed ion
               Chromatograms shall be normalized to the largest nonsolvent
               component and shall contain the following header information:

               •   EPA sample number,

               •   Date and time of analysis,

               •   GC/MS instrument identifier,

               •   Lab file identifier, and

               •   Analyst ID.

2.6.4.2.3.1       Internal standards and surrogate compounds shall be labeled
                  with the names of compounds, either directly out from the
                  peak or on a printout of retention times if retention times
                  are printed over the peak.

2.6.4.2.3.2       If automated data system procedures are used for preliminary
                  identification and/or quantitation of the target compounds,
                  the complete data system report shall be included in all
                  sample data packages, in addition to the reconstructed ion
                  chromatogram.  The complete data system report shall include
                  all of the information listed below.   For laboratories which
                  do not use the automated data system procedures, a
                  laboratory "raw data sheet" containing the following
                  information shall be included in the sample data package, in
                  addition to the chromatogram.

                  •   EPA sample number,

                  •   Date and time of analysis,
                                     B-19                              OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
Semivolatiles
                  •   Retention time or scan number of identified target
                      compounds,

                  •   Ion used for quantitation with measured area,

                  •   Copy of area table from data system,

                  •   GC/MS instrument identifier, and

                  •   Lab file identifier.

2.6.4.2.3.3       In all instances where the data system report has been
                  edited, or where manual integration or quantitation has been
                  performed, the GC/MS operator shall identify such edits or
                  manual procedures by initialing and dating the changes made
                  to the report, and shall  include the integration scan range.
                  In addition, a hardcopy printout of the EICP of the
                  quantitation ion displaying the manual integration shall be
                  included in the raw data.   This applies to all compounds
                  listed in Exhibit C (Semivolatiles), internal standards and
                  system monitoring compounds.

                  •   EICPs displaying each  manual integration.

2.6.4.2.4      Other Required Information.   For each sample, by each compound
               identified, the following shall be included in the data
               package.

               •   Copies of raw spectra and copies of background-subtracted
                   mass spectra of target compounds listed in Exhibit C
                   (Semivolatiles) that are  identified in the sample and
                   corresponding background-subtracted target compound
                   standard mass spectra.  Spectra shall be labeled with EPA
                   sample number, lab file  identifier, and date and time of
                   analysis, and GC/MS instrument identifier compound names
                   shall be clearly marked  on all spectra.

               •   Copies of mass spectra of non-surrogate/non-internal
                   standard organic compounds not listed in Exhibit C
                   (Volatiles and Semivolatiles) with associated best-match
                   spectra (maximum of three best matches).  This includes the
                   mass spectra for tentatively identified alkanes.  Spectra
                   shall be labeled with EPA sample number, lab file
                   identifier, and date and  time of analysis, and GC/MS
                   instrument identifier compound names shall be clearly
                   marked on all spectra.

2.6.4.3     Semivolatiles Standards Data

2.6.4.3.1      Initial calibration data (Form VI SV-1, SV-2) shall be included
               in order by instrument, if more than one instrument used.

               •   Semivolatile standard(s)  reconstructed ion chromatograms
                   and quantitation reports  for the initial  (five-point)
                                     B-20                              OLM03.1

-------
                                                          Exhibit B—Section  2
                          Reporting  Requirements and Order of Data Deliverables
                                                                 Semivolatiles
                    calibration,  labeled  as  in Section 2.6.4.2.3.  Spectra are
                    not  required.

                •    All  initial calibration  data that pertain to samples  in the
                    data package  shall be included, regardless of when  it was
                    performed and  for which  Case.  When more than one initial
                    calibration is performed, the data shall be in
                    chronological  order,  by  instrument.

                •    EICPs displaying each manual integration.

2.6.4.3.2       Continuing calibration data  (Form VII SV-1, SV-2) shall be
                included in order by instrument, if more than one instrument
                used.

                •    Semivolatile standard(s) reconstructed ion chromatograms
                    and  quantitation reports for all continuing (12-hour)
                    calibrations,  labeled as in Section 2.6.4.2.3.  Spectra are
                    not  required.

                •    When more than one continuing calibration is performed,
                    forms shall be in chronological order, by instrument.

                •    EICPs displaying each manual integration.

2.6.4.3.3       In all instances where the data system report has been edited,
               or where manual integration or quantitation has been performed,
               the GC/MS operator shall  identify such edits or manual
               procedures by initialing  and dating the changes made to the
               report,  and shall include the integration scan range.  In
               addition, a hardcopy printout of the EICP of the quantitation
               ion displaying the manual integration shall be included in the
               raw data.  This applies to all compounds listed in Exhibit C
                (Semivolatiles),  internal standards and system monitoring
               compounds.

2.6.4.4     Semivolatiles Raw QC Data

2.6.4.4.1      DFTPP data shall  be arranged in chronological order by
               instrument for each 12-hour period,  for each GC/MS system
               utilized.

               •   Bar graph spectrum,  labeled as in Section 2.6.4.2.3.

               •   Mass listing,  labeled as in Section 2.6.4.2.3.

               •   Reconstructed total ion chromatogram,  labeled as in Section
                   2.6.4.2.3.
                                     B-21                              OLM03.0

-------
 Exhibit  B—Section 2
 Reporting Requirements  and  Order  of Data  Deliverables
 Semivolatiles
 2.6.4.4.2       Blank  data  shall be  included  in  chronological order  by
                extraction  date.

                NOTE:   This order  is different from that used for  samples.

                •   Tabulated  results  (Form I SV-1, SV-2).

                •   Tentatively identified compounds  (Form  I SV-TIC) even if
                    none  are found.

                •   Reconstructed  ion  chromatogram(s) and quantitation
                    report(s), labeled as in  Section  2.6.4.2.3.

                •   Target  compound spectra with laboratory-generated standard,
                    labeled as in  Section 2.6.4.2.4.  Data  systems which are
                    incapable  of dual display shall provide spectra  in the
                    following  order:

                       Raw target compound spectra.

                       Enhanced or background-subtracted spectra.

                       Laboratory-generated standard spectra.

                •   GC/MS library  search spectra for tentatively identified
                    compounds,  labeled as in Section 2.6.4.2.4.

                •   Quantitation/calculation of tentatively identified compound
                    concentrations.

2.6.4.4.3       Semivolatiles Matrix Spike Data

                •   Tabulated results  (Form I SV-1, SV-2) of target  compounds.
                    Form  I  SV-TIC  is not required.

                •   Reconstructed  ion chromatogram(s)  and quantitation
                    report(s),  labeled as in Section 2.6.4.2.3.  Spectra are
                    not required.

2.6.4.4.4       Semivolatiles Matrix Spike Duplicate Data

                •    Tabulated results (Form I SV-1, SV-2) of target  compounds.
                    Form  I  SV-TIC  is not required.

                •    Reconstructed  ion chromatogram(s)  and quantitation
                    report(s),  labeled as in Section 2.6.4.2.3.  Spectra are
                    not required.

2.6.4.4.5       Semivolatile GPC Data.   The UV traces for the GPC calibration
                solution and the reconstructed ion chromatogram and  data system
                reports for the GPC blank shall be arranged in chronological
                order by GPC for the GPC calibration.
                                     B-22                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                           Pesticides/Aroclors
               •   UV traces labeled with the GPC column identifier, date of
                   calibration, and compound names.   Compound names shall be
                   placed directly out from the peak, or on the printout of
                   retention times when the retention times are printed
                   directly over the peak.

               •   Reconstructed ion chromatogram and data system report(s)
                   labeled as specified in Section 2.6.4.2.3 for GPC blank
                   analysis.

               •   Reconstructed ion chromatogram and data system report(s)
                   for all standards used to quantify compounds in the GPC
                   blank labeled as specified in Section 2.6.4.2.3 (continuing
                   calibration standard).

2.6.5    Pesticide/Aroclor Data

2.6.5.1     Pesticide/Aroclor QC Summary

2.6.5.1.1      Surrogate Percent Recovery Summary (Form II PEST).

2.6.5.1.2      Matrix Spike/Matrix Spike Duplicate Summary (Form III PEST).

2.6.5.1.3      Method Blank Summary (Form IV PEST):   If more than a single
               form is necessary,  forms shall be arranged in chronological
               c  der by date of analysis of the Llank.

2.6.5.2     Pesticide/Aroclor Sample Data.  Sample data shall be arranged in
            packets with the Organic Analysis Data Sheet (Form I PEST),
            followed by the raw data for pesticide samples.  These sample
            packets should then be placed in increasing EPA sample number
            order,  considering both letters and numbers.

2.6.5.2.1      Target Compound Results, Organic Analysis Data Sheet (Form I
               PEST).  Tabulated results (identification and quantitation) of
               the specified target compounds (Exhibit C, Pesticides/Aroclors)
               shall be included.   The validation and release of these results
               is authorized by a specific, signed statement in the SDG
               Narrative (see Section 2.6.1).  In the event that the
               laboratory manager cannot verify all  data reported for each
               sample, the laboratory manager shall  provide a detailed
               description of the problems associated with the sample in the
               SDG Narrative.

2.6.5.2.2      Copies of Pesticide Chromatograms.  Positively identified
               compounds shall be labeled with the names of compounds, either
               directly out from the peak on the chromatogram, or on a
               printout of retention times on the data system printout if
               retention times are printed over the  peak on the chromatogram.
               All chromatograms shall meet the acceptance criteria in Exhibit
               D PEST, and shall be labeled with the following information:
                                     B-23                              OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
Pesticides/Aroclors
               •   EPA sample number,

               •   Volume injected

               •   Date and time of injection,

               •   GC column identifier (by stationary phase and internal
                   diameter),

               •   GC instrument identifier,  and

               •   Scaling factor.

2.6.5.2.3      Copies of pesticide chromatograms from second GC column shall
               be included and labeled as in  Section 2.6.5.2.2.

2.6.5.2.4      Data System Printout.   A printout of retention time and
               corresponding peak height or peak area shall accompany each
               chromatogram.  The printout shall be labeled with the EPA
               sample number.   In all  instances where the data system report
               has been edited,  or where manual integration or quantitation
               has been performed, the GC/EC  operator must identify such edits
               or manual procedures by initialing and dating the changes made
               to the report,  and shall include the integration time range.

2.6.5.2.5      All manual work sheets  shall be  included in the sample data
               package.

2.6.5.2.6      Other Required Information.  If  pesticides/Aroclors are
               confirmed by GC/MS, the Contractor shall submit copies of
               reconstructed ion chromatograms, raw spectra and background-
               subtracted mass spectra of target compounds listed in Exhibit C
               (Pesticides/Aroclors)  that are identified in the sample and
               corresponding background-subtracted TCL standard mass spectra.
               Compound names shall be clearly  marked on all spectra.  For
               multicomponent pesticides/Aroclors confirmed by GC/MS, the
               Contractor shall submit mass spectra of three major peaks of
               im1 Iticomponent compounds from  samples and standards.

2.6.5.3     Pesticide/Aroclor Standards Data

2.6.5.3.1      Initial Calibration of  Single  Component Analytes (Form VI PEST-
               1, PEST-2):  for all GC columns, all instruments, in
               chronological order by  GC column and instrument.

2.6.5.3.2      Initial Calibration of  Multicomponent Analytes  (Form VI PEST-
               3):  for all GC columns, all instruments, in chronological
               order by GC column and  instrument.

2.6.5.3.3      Analyte Resolution Summary (Form VI PEST-4):  for all GC
               columns and instruments, in chronological order by GC column
               and instrument.
                                     B-24                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                           Pesticides/Aroclors
2.6.5.3.4      Performance Evaluation Mixture  (Form VI PEST-5):  for all GC
               columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.5      Individual Standard Mixture A (Form VI PEST-6):  for all GC
               columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.6      Individual Standard Mixture B (Form VI PEST-7):  for all GC
               columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.7      Calibration Verification Summary (Form VII PEST-1):  for all
               performance evaluation mixtures and instrument blanks, on all
               GC columns and instruments, in chronological order by GC column
               and instrument.

2.6.5.3.8      Calibration Verification Summary (Form VII PEST-2):  for all
               mid-point concentrations of Individual Standard Mixtures A and
               B and instrument blanks used for calibration verification, on
               all GC columns and instruments,  in chronological order by GC
               column and instrument.

2.6.5.3.9      Analytical Sequence (Form VIII PEST):   for all GC columns and
               instruments,  in chronological order by GC column and
               instrument.

2.6.5.3.10     Florisil Cartridge Check (Form IX PEST-1):  for all lots of
               cartridges used to process samples in the SDG.

2.6.5.3.11     Pesticide GPC Calibration (Form IX PEST-2):   for all GPC
               columns, in chronological order by calibration date.

2.6.5.3.12     Pesticide Identification Summary for Single Component Analytes
               (Form X PEST-1):   for all samples with positively identified
               single component analytes,  in order by increasing EPA sample
               number.

2.6.5.3.13     Pesticide Identification Summary for Multicomponent Analytes
               (Form X PEST-2):   for all samples with positively identified
               multicomponent analytes, in order by increasing EPA sample
               number.

2.6.5.3.14     Chromatograms and data system printouts shall be included for
               all standards including the following:

               •   Resolution check mixture.

               •   Performance evaluation mixtures,  all.

               •   Individual Standard Mixture A,  at  three concentrations,
                   each initial calibration.
                                     B-25                              OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
Pesticides/Aroclors
               •   Individual Standard Mixture B,  at three concentrations,
                   each initial calibration.

               •   All multicomponent analytes (toxaphene and Aroclors), each
                   initial calibration.

               •   All mid-point concentrations of Individual Standard
                   Mixtures A and B used for calibration verification.

               •   All multicomponent analyte standards analyzed for
                   confirmation.

2.6.5.3.15     A printout of retention time and corresponding peak height or
               peak area shall accompany each chromatogram.  The printout
               shall be labeled with the EPA sample number.  In addition, all
               chromatograms shall meet the acceptance criteria in Exhibit D
               PEST, and shall be labeled with the following:

               •   EPA sample number for the standard, e.g.,  INDA1, INDA2,
                   etc.  (See Section 4 for details.)

               •   Label all standard peaks for all individual compounds
                   either directly out from the peak on the chromatogram or on
                   the printout of retention times on the data system printout
                   if retention times are printed  over ti e peak on the
                   chromatogram.

               •   Total nanograms injected for each standard.  When total
                   nanograms injected appear on the printout, it is not
                   necessary to include them on the chromatogram.

               •   Date and time of injection.

               •   GC column identifier (by stationary phase and internal
                   diameter).

               •   GC instrument identifier.

               •   Scaling factor.

               NOTE: In all instances where the data system report has been
               edited, or where manual integration or quantitation has been
               performed, the GC/EC operator must  identify such edits or
               manual procedures by initialing and dating the changes made to
               the report, and shall include the integration time range.

2.6.5.4     Pesticide/Aroclor Raw QC Data

2.6.5.4.1      Blank data shall be arranged by type of blank  (method,
               instrument, sulfur cleanup) and shall be in chronological order
               by instrument.

               NOTE:  This order is different from that used  for samples.

                                     B-26                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables
                                                           Pesticides/Aroclors
               •   Tabulated results (Form I PEST).

               •   Chromatogram(s)  and data system printout(s) (GC) for each
                   GC column and instrument used for analysis, labeled as in
                   Sections 2.6.5.2.2 and 2.6.5.2.4.

2.6.5.4.2      Matrix Spike Data

               •   Tabulated results (Form I PEST) of target compounds.

               •   Chromatogram(s)  and data system printout(s) (GC),  labeled
                   as in Sections 2.6.5.2.2 through 2.6.5.2.4.

2.6.5.4.3      Matrix Spike Duplicate Data

               •   Tabulated results (Form I PEST) of target compounds.

               •   Chromatogram(s)  and data system printout(s) (GC),  labeled
                   as in Sections 2.6.5.2.2 through 2.6.5.2.4.

2.6.5.5     Raw GPC Data

2.6.5.5.1      GPC Calibration.  The UV traces for the GPC calibration
               solution, chromatograms, and the data system reports for the
               GPC blank shall be arranged in chronological order for the GPC
               calibration.

               •   UV traces labeled with the GPC column identifier,  date of
                   calibration, and compound names.  Compound names shall be
                   placed directly out from the peak, or on the printout of
                   retention times when the retention times are printed
                   directly over the peak.

               •   Chromatogram and data system report(s) labeled as specified
                   in Sections 2.6.5.2.2 and 2.6.5.2.4 for GPC blank analysis.

               •   Chromatogram and data system report(s) for all standards
                   used to quantify compounds in the GPC blank labeled as
                   specified in Section 2.6.5.3.15  (i.e., Individual Standard
                   Mixture A, Individual Standard Mixture B, and the
                   Aroclor/toxaphene standards).

2.6.5.5.2      GPC Calibration Check.  The Chromatogram and the data system
               report(s) shall be arranged in chronological order for the GPC
               calibration check.

               •   Chromatograms and data system printouts labeled as
                   specified in Sections 2.6.5.2.2 and 2.6.5.2.4 for the GPC
                   calibration check solution analyses.

               •   Chromatogram and data system report(s) for standards used
                   to quantify compounds in the GPC calibration check solution
                   or used to assess the Aroclor pattern labeled as specified

                                     B-27                              OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables
Pesticides/Aroclors
                   in Section 2.6.5.3.15 (i.e.,  Individual Standard Mixtures A
                   and B and Aroclor Standard Mixture 1016/1260 form the
                   initial calibration sequence).

2.6.5.6     Raw Florisil Data.  The chromatogram and data system report(s)
            shall be arranged in chronological order by Florisil cartridge
            performance check analyses.

            •   Chromatograms and data system reports labeled as specified in
                Sections 2.6.5.2.2 and 2.6.5.2.4 for the florisil cartridge
                performance check analyses.

            •   Chromatograms and data system reports for standard analyses
                used to quantify compounds in the Florisil cartridge
                performance check analysis,  labeled as specified in Section
                2.6.5.3.15 (i.e., Individual Standard Mixture A and Individual
                Standard Mixture B and the 2,4,5 Trichlorophenol solution).

2.7   Complete SDG File.  As specified in Section 1, the Contractor shall
      deliver one Complete SDG File (CSF) including the original sample data
      package to the Region concurrently with delivery of the sample data
      package to SMO.  (Delivery to EMSL/LV is only required upon written
      request.)

2.7.1    The CSF will contain all original documents specified in Sections 3
         and 4  .id in Form DC-2 (see Section 4).  No photocopies of original
         documents will be placed in the CSF unless the original data was
         initially written in a bound notebook,  maintained by the Contractor,
         or the originals were previously submitted to the Agency with another
         Ca.se/SDG in accordance with the requirements described in Exhibit F.
         The contents of the CSF shall be numbered according to the
         specifications described in Section 3.20.

2.7.2    The CSF will consist of the following original documents in addition
         to the documents in the sample data package.

         NOTE:  All SDG-related documentation may be used or admitted as
         evidence in subsequent legal proceedings.  Any other SDG-specific
         documents generated after the CSF is sent to EPA, as well as copies
         that are altered in any fashion, are also deliverables to EPA.
         (Deliver the original to the Region and a copy to SMO.  Delivery to
         EMSL/LV is only upon written request.)

2.7.2.1     The original sample data package.

2.7.2.2     A completed and signed document inventory sheet  (Form DC-2).

2.7.2.3     All original shipping documents including, but not limited to, the
            following documents:

            •   EPA Chain-of-Custody Record,
            •   Airbills,
            •   EPA Traffic Reports, and

                                     B-28                              OLM03.0

-------
                                                          Exhibit B—Section 2
                         Reporting Requirements and Order of Data Deliverables


            •   Sample tags (if present) sealed in plastic bags.

2.7.2.4     All original receiving documents including, but not limited to,
            the following documents:

            •   Form DC-1,

            •   Other receiving forms or copies of receiving logbooks, and

            •   SDG cover sheet.

2.7.2.5     All original laboratory records, not already submitted in the
            sample data package, of sample transfer, preparation and analysis
            including, but not limited to, the following documents:

            •   Original preparation and analysis forms or copies of
                preparation and analysis logbook pages,

            •   Internal sample and sample extract transfer chain-of-custody
                records,

            •   Screening records, and

            •   All instrument output, including strip charts from screening
                activities.

2.7.2.6     All other original SDG-specific documents in the possession of the
            Contractor including,  but not limited to, the following documents:

            •   Telephone contact logs,

            •   Copies of personal logbook pages,

            •   All hand-written SDG-specific notes, and

            •   Any other SDG-specific documents not covered by the above.

2.7.3    If the Contractor does submit SDG-specific documents to EPA after
         submission of the CSF, the documents should be identifier with unique
         accountable numbers,  a revised Form DC-2 should be submitted, and the
         unique accountable numbers and the locations of the documents in the
         CSF should be recorded in the "Other Records" section on the revised
         Form DC-2.  Alternatively, the Contractor may number the newly
         submitted SDG-specific documents to EPA as a new CSF and submit a new
         Form DC-2.  The revised Form DC-2 or new Form DC-2 should be
         submitted to the EPA Regions only.

2.8   Data in Computer-Readable Form.  The Contractor shall provide a
      computer-readable copy of the data on data reporting Forms I-X for all
      samples in the SDG as specified in Exhibit H, and delivered as specified
      in the Contract Schedule (Contract Performance/Delivery Schedule).
      Computer-readable data deliverables shall be submitted on IBM or
      IBM-compatible, 5.25-inch floppy double-sided, double-density 360 K-byte


                                     B-29                              OLM03.0

-------
Exhibit B—Section 2
Reporting Requirements and Order of Data Deliverables


      or a high density 1.2 M-byte, or a 3.5-inch double-sided double-density
      720 K-byte or 1.44 M-byte diskette.

2.8.1    When submitted, diskettes shall be packaged and shipped in such a
         manner that the diskette(s) cannot be bent or folded, and will not be
         exposed to extreme heat or cold or any type of electromagnetic
         radiation.  The diskette(s) shall be included in the same shipment as
         the hardcopy data and shall, at a minimum, be enclosed in a diskette
         mailer.  The diskettes shall be labeled as specified in Exhibit H,
         Section 8.4.

2.8.2    The data shall be recorded in ASCII, text file format, and shall
         adhere to the file, record, and field specifications listed in
         Exhibit H.

2.9   GC/MS Tapes.   The Contractor shall adhere to the requirements in Exhibit
      E.

2.10  Extracts.   The Contractor shall preserve sample extracts at less than 4°
      C but not greater than 6° C in bottles/vials with Teflon-lined septa.
      Extract bottles/vials shall be labeled with EPA sample number,  Case
      number and SDG number.  The Contractor shall maintain a logbook of
      stored extracts,  listing EPA sample numbers and associated Case and SDG
      numbers.   The Contractor shall retain extracts for 365 days following
      submission of the reconciled complete sample data package.   During that
      time,  the Contractor shall submit extracts and associated logbook pages
      within seven  days following receipt of a written request from the
      Administrative Project Officer,  Technical Project Officer,  or EMSL/LV.
                                     B-30                               OLM03.0

-------
                                                          Exhibit B—Section 3
                                                            Forms Instructions
                                                           General Information
3.0   FORMS INSTRUCTIONS

3.1   Introduction.  This section includes specific instructions for
      completing the data reporting forms required under this contract.  Each
      of the forms is specific to a given fraction (volatile, semivolatile, or
      pesticide/Aroclor) and, in some instances, specific to a given matrix
      (water or soil) within each fraction.  The Contractor shall submit only
      those forms pertaining to the fractions analyzed for a given sample(s).
      For instance, if a sample is scheduled for volatiles analysis only, the
      Contractor shall provide only forms for the volatile fraction.  NOTE:
      There are two pages relating to the semivolatile fraction for Forms I,
      VI, VII, and VIII.  Whenever semivolatiles are analyzed and one of these
      forms is required, both pages (SV-1 and SV-2) shall be submitted.

3.2   General Information.  The Contractor shall report values on the hardcopy
      forms according to the individual form instructions in this section.
      For example, results for concentrations of volatile target compounds
      shall be reported to two significant figures if the value is greater
      than or equal to 10.  Values that exceed the maximum length allowed
      shall be reported to the maximum possible, maintaining the specified
      decimal place.

3.2.1    The data reporting forms presented in Section 4 have been designed in
         conjunction with the computer-readable data format specified in
         Exhibit H.  The specific length of each variable for
         computer-readable data transmission purposes is also given in Exhibit
         H.   Information entered on these forms shall not exceed the size of
         the field given on the form,  including such laboratory-generated
         items as lab name and lab sample identifier.

         NOTE:   The space provided for entries on the hardcopy forms (Section
         4)  is greater in some instances than the length prescribed for the
         variable as written to diskette (see Exhibit H).   Greater space is
         provided on the hardcopy forms for visual clarity.

3.2.2    When submitting data,  the Contractor shall reproduce all characters
         that appear on the data reporting forms in Section 4.   The format of
         the forms submitted shall be identical to that shown in the contract.
         No information may be added,  deleted,  or moved from its specified
         position without prior written approval of the Administrative Project
         Officer.   The names of the various fields and compounds (e.g., "Lab
         Code,"  "Chloromethane")  shall appear as they do on the forms in the
         contract,  including the options specified in the form (e.g., "Matrix:
         (soil/water)" shall appear,  not just "Matrix").   For items appearing
         on the uncompleted forms (Section 4),  the use of uppercase and
         lowercase letters is optional.

3.2.3    Alphabetical entries made on the forms by the Contractor shall be in
         ALL UPPERCASE letters (e.g.,  "LOW",  not "Low" or "low").   If an entry
         does not fill the entire blank space provided on the form,  null
         characters shall be used to remove the remaining underscores that
         comprise the blank line.   See Exhibit H for more detailed
                                     B-31                              OLM03.0

-------
 Exhibit  B—Section  3
 Forms  Instructions
 General  Information
          instructions.  However, the Contractor shall not remove the
          underscores or vertical bars that delineate "boxes" on the forms.
          The only exception would be those underscores at the bottom of a
          "box" that are intended as a data entry line.   (For instance, on Form
          2A, line 30, if data is entered on line 30, it will replace the
          underscores.)

3.3   Header Information.  Six pieces of information are common to the header
      section of each data reporting form:  lab name, contract, lab code, Case
      number, SAS number and SDG number.  Except as noted for SAS number, this
      information shall be entered on every form and shall match on every
      form.

3.3.1     Lab Name.  The lab name shall be the name chosen by the Contractor to
          identify the laboratory.  It shall not exceed 25 characters.

3.3.2     Contract.  Contract refers to the number of the EPA contract under
          which the analyses were performed.

3.3.3     Lab Code.  The lab code is an alphabetical abbreviation of up to six
          letters, as assigned by EPA, to identify the laboratory and aid in
          data processing.  This lab code will be assigned by EPA at the time a
          contract is awarded, and shall not be modified by the Contractor,
          except at the direction of EPA.  If a change of name or ownership
          occurs at the laboratory,  the lab code will remain the same until the
         Contractor is directed by EPA to use another lab code.

3.3.4     Case Number.   The Case number is the EPA-assigned Case number
          associated with the sample.   This number is reported on the Traffic
         Report.

3.3.5     SAS Number.  The SAS number is the EPA-assigned number for analyses
         performed under Special Analytical Services.  If samples are to be
         analyzed under SAS only and reported on these forms, then enter the
         SAS number and leave the case number blank.  If samples are analyzed
         according to the Routine Analytical Services (RAS) protocols and have
         additional SAS requirements, list both the Case number and the SAS
         number on all forms.  If there are no SAS requirements, leave the
          "SAS No." field blank.

         NOTE:   Some samples in an SDG may have a SAS number while others may
         not.

3.3.6    SDG Number.  The "SDG No." field is for the sample delivery group
         number.   It is the EPA sample number of the first sample received in
         the SDG.  When several samples are received together in the first SDG
         shipment,  the SDG number shall be the lowest sample number
          (considering both alpha and numeric designations) in the first group
         of samples received under the SDG.

3.3.7    Sample Number.   This number appears either in the upper righthand
         corner of the form,  or as the left column of a table summarizing data
                                     B-32                              OLM03.0

-------
                                                          Exhibit B—Section 3
                                                            Forms Instructions
                                                           General Information
          from  a number of  samples.  When the EPA sample number is entered  in
          the triple-spaced box  in the upper righthand corner of Form  I, Form
          IV, or Form X,  it should be entered on the middle line of the three
          lines that comprise the box.

3.3.7.1     The Contractor shall identify all samples, including dilutions and
            reanalyses,  matrix  spikes, matrix spike duplicates, blanks, and
            standards with an EPA sample number.  For field samples,  matrix
            spikes and matrix spike duplicates, the EPA sample number is the
            unique identifying  number given in the Traffic Report that
            accompanied  that sample.  In order to facilitate data assessment,
            the Contractor shall use the following sample suffixes:

            XXXXX      =   EPA sample number
            XXXXXMS    =   Matrix spike sample
            XXXXXMSD   =   Matrix spike duplicate  sample
            XXXXXRE    =   Re-extracted and  reanalyzed  sample
            XXXXXDL    =   The suffix DL is  appended  to the EPA
                          sample number to  indicate  that  the
                          analytical results  are  a  result of  a
                          dilution  of  the original  analysis
                          (reported  as EPA  sample XXXXX).   See
                          Exhibit D  for requirements  for
                          dilutions.

3.3.7.2     There may be instances when all samples analyzed must be  listed on
            the form, regardless of whether or not they are part of the SDG
            being reported (e.g., Form VIII PEST).   In these instances, use
            ZZZZZ as the EPA sample number for any sample analysis not
            associated with the SDG being reported.

3.3.7.3     For blanks,  the Contractor shall use the following identification
            scheme for the EPA  sample number:

            •   Volatile method blanks shall be identified as VBLK##.

            •   Volatile instrument blanks shall be identified as VIBLK##.

            •   Volatile storage blanks shall be identified as VHBLK##.

            •   Semivolatile method blanks shall be identified as SBLK##.

            •   Pesticide/Aroclor method blanks and/or sulfur cleanup blanks
                shall be identified as PBLK##.

            •   Pesticide/Aroclor instrument blanks shall be identified as
                PIBLK##.

3.3.7.4     The EPA sample number shall be unique for each blank within an
            SDG.   Within a fraction, the Contractor shall achieve this by
            replacing the two-character terminator (##)  of the identifier with

                                     B-33                              OLM03.0

-------
Exhibit B—Section 3
Forms Instructions
General Information
3.3.7.5
3.3.7.6
one or two characters or numbers, or a combination of both.  For
example, possible identifiers for volatile blanks would be VBLK1,
VBLK2, VBLKA1, VBLKB2, VBLK10, VBLKAB, etc.  If the method blank
is analyzed on multiple instruments, then an additional two-
character suffix shall be added to make the blank EPA sample
number unique.
Volatile and semivolatile standards shall be identified as
FSTD***##, where
   F    =   Fraction code (V for volatiles; S for semivolatiles).
   STD  =   Standard.
               * * *  =
               ##
            Concentration of volatile standards in ug/L (e.g.,
            010, 020, 050, 100,  and 200)  or the amount injected in
            ng for semivolatile  standards (e.g., 020, 050, 080,
            120, and 160).
            One or two characters,  numbers,  or combinations of
            both to create a unique EPA sample number within an
            SDG.
The Contractor shall use the following scheme to identify
pesticide/Aroclor standards:
                           Name
                                   EPA Sample Number
             Individual Mix A  (low point)
             Individual Mix A  (mid-point)
             Individual Mix A  (high point)
             Individual Mix B  (low point)
             Individual Mix B  (mid-point)
             Individual Mix B  (high point)
             Resolution Check
             Performance Evaluation Mixture
             Toxaphene
             Aroclor  1016
             Aroclor  1221
             Aroclor  1232
             Aroclor  1242
             Aroclor  1248
             Aroclor  1254
             Aroclor  1260
             Aroclor  1016/1260 Mix
                                        INDAL##
                                        INDAM##
                                        INDAH##
                                        INDBL##
                                        INDBM##
                                        INDBH##
                                        RESC##
                                         PEM##
                                       TOXAPH##
                                       AR1016##
                                       AR1221##
                                       AR1232##
                                       AR1242##
                                       AR1248##
                                       AR1254##
                                       AR1260##
                                       AR1660##
                                     B-34
                                                           OLM03.0

-------
                                                          Exhibit B—Section 3
                                                            Forms Instructions
                                                           General Information
            The Contractor shall replace the two-character terminator  (##) of
            the identifier with one or two characters or numbers, or a
            combination of both, to create a unique EPA sample number within
            an SDG.

3.3.7.7     If the standards are injected onto both GC columns on the same
            instrument simultaneously, the same EPA sample number may be used
            for reporting data for the standards for both columns.  If
            simultaneous injections are not made, then the same number shall
            not be used.

3.3.7.8     The EPA sample number for GPC shall be GPC#########, where
            ######### is the GPC column ID.  If the GPC column ID is more than
            nine characters, truncate at the ninth character.

3.3.7.9     The EPA sample number for florisil shall be FLO#########, where
            ######### is the florisil cartridge lot number.  If the florisil
            cartridge lot number is more than nine characters,  truncate at the
            ninth character.

3.3.8    Other Common Fields.  Several other pieces of information are common
         to many of the data reporting forms.   These include matrix, sample
         weight/volume,  level,  lab sample identifier,  and lab file identifier.

         •   In the "Matrix" field,  enter SOIL for soil/sediment samples and
             WATER for water samples.

             NOTE:   The matrix  shall be spelled out.   Abbreviations such as S
             or W shall not be  used.

         •   In the "Sample wt/vol"  field,  enter the number of  grams (for
             soil)  or milliliters (for water)  of sample used in the first
             blank.   Enter the  units,  either G or ML,  in the second blank.

         •   The "Level"  field  is used for the volatile and semivolatile
             fractions.   Enter  the determination of concentration level made
             from the screening of soils.   Enter as LOW or MED,  not L or M.
             All water samples  shall  be entered as LOW.

             NOTE:   There is no differentiation between low and medium soil
             samples for  the pesticide/Aroclor fraction,  and no level is
             entered on  any of  these  forms.

         •   The lab sample identifier is  a unique laboratory-generated
             internal identifier  pertaining to a particular analysis.   The
             Contractor  can enter up  to 12  alpha-numeric  characters in the
             "Lab  Sample  ID"  field.   The Contractor may use the EPA sample
             number  ?.s the  lab  sample  identifier.

         •   The lab file  identifier  is the unique laboratory-generated name
             of  the  GC/MS  data  system  file  containing  information pertaining
                                     B-35                               OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form I
             to a particular analysis.  The Contractor can enter up to 14
             alpha-numeric characters in the "Lab File ID" field.

3.3.8.1     The "Instrument ID" field is common to the forms containing
            calibration data.  The identifier used by the Contractor shall
            include some indication of the manufacturer and/or model of the
            instrument, and shall contain additional characters that
            differentiate between all instruments of the same type in the
            laboratory.

3.3.8.2     Forms II, IV, V, VIII, IX, and X contain a field labeled "page _
            of _" in the bottom lefthand corner.  If the number of entries
            reg_uired on any of these forms exceeds the available space,
            continue entries on another copy of the same fraction-specific
            form, duplicating all header information.   If a second page is
            required, number the pages consecutively (i.e., "page I of 2" and
            "page 2 of 2").  If a second page is not required, number the page
            "page 1 of 1."

            NOTE:  These forms are fraction-specific,  and often matrix-
            specific within a fraction.   For example,  Form II VOA-1 and Form
            II VOA-2 are for different data.  Therefore,  do not number the
            pages of all six versions of Form II as "1 of 6," "2 of 6," etc.
            Number only pages corresponding to the fraction-specific and
            matrix-specific form.

3.3.9    Rounding Rule.  For rounding off numbers to the appropriate level of
         precision,  the Contractor shall follow these rules.   If the figure
         following those to be retained is less than 5,  drop it (round down).
         If the figure is greater than 5,  drop it and increase the last digit
         to be retained by 1 (round up).  If the figure following the last
         digit to be retained equals 5,  round up if the digit to be retained
         is odd,  and round down if that digit is even.

3.4   Organic Analysis Data Sheet (Form I, All Fractions)

3.4.1    Purpose.  This form is used for tabulating and reporting sample
         analysis,  including blank,  matrix spike, and matrix spike duplicate
         results for target compounds.  If all fractions are not requested for
         analysis,  only the pages for the fractions required shall be
         submitted.   For example,  if only volatiles analysis is requested,
         Form I VOA and Form I VOA-TIC shall be submitted.  If only the
         pesticide/Aroclor fraction is requested for analysis, Form I PEST
         shall be submitted.  Furthermore, pesticide instrument blanks
         (PIBLKs) shall be reported on a per column/per analysis basis on Form
         I PEST.   Each PIBLK shall be named with a unique EPA sample number.

3.4.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.
                                     B-36                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                        Form I
3.4.2.1     For soil  samples analyzed for volatiles, enter the non-decanted
            percent moisture in the "% Moisture: not dec." field on Form I
            VOA.  This  is the only percent moisture determination made for
            volatiles since the entire contents of the VOA vial are considered
            as the sample.  For water samples, leave this field blank.

3.4.2.2     For soil  samples analyzed for semivolatiles and
            pesticides/Aroclors, enter the values for the percent moisture
            determined  during the analysis in the "% Moisture" field on Form I
            SV-1, SV-2  or PEST.  In the "decanted (Y/N)" field, enter Y if the
            sample had  standing water above the soil/sediment that was
            decanted, or N if no water was decanted off the surface of the
            sample.  Report percent moisture (decanted or not decanted) to the
            nearest whole percentage point (e.g., 5%, not 5.3%).  For water
            samples, method blanks, sulfur cleanup blanks and instrument
            blanks, leave these fields on Form I blank.

3.4.2.3     For volatiles, enter the GC column identifier in the "GC Column"
            field on Form I VOA, and the internal diameter in millimeters
            (mm), to two decimal places, in the "ID" field.  For packed
            columns, convert the internal diameter rrom inches to millimeters
            as necessary before entering in the "ID" field.

3.4.2.4     For pesticides/Aroclors, enter the method of extraction in the
            "Extraction" field on Form I PEST as SEPF fcr separatory funnel,
            CONT for continuous liquid-liquid extraction, or SONC for
            sonication  (soils only).

3.4.2.5     If gel permeation chromatography (GPC) was performed, enter Y in
            the "GPC Cleanup" field on Form I SV-1,  SV-2 or PEST.  Enter N in
            this field  if GPC was not performed.

            NOTE:  GPC  is required for all soil samples analyzed for
            semivolatiles and pesticides/Aroclors; therefore, all forms for
            soil samples will contain a Y in this field.

3.4.2.6     For soil samples only,  enter the pH for semivolatiles and
            pesticides/Aroclors, reported to 0.1 pH units,  on Form I SV-1, SV-
            2 or PEST.

3.4.2.7     Enter the date of sample receipt at the laboratory, as noted on
            the Traffic Report (i.e.,  the VTSR), in the "Date Received" field.
            The date shall be entered as MM/DD/YY.

3.4.2.8     Complete the "Date Extracted" and "Date Analyzed" fields in the
            same format (MM/DD/YY).  When continuous liquid-liquid extraction
            procedures  are used for water samples, enter the date that the
            procedure was started in the "Date Extracted" field.  If
            separatory  funnel (pesticides only) or sonication procedures are
            used, enter the date that the procedure was completed in the "Date
            Extracted"  field.  For pesticide/Aroclor samples, enter the date
            of the first GC analysis performed in the "Date Analyzed" field.
                                     B-37                              OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form I
            The date of sample receipt will be compared with the extraction
            and analysis dates of each fraction to ensure that contract
            holding times were not exceeded.

3.4.2.9     If a medium soil sample is analyzed for volatiles, enter total
            volume of the methanol extract in microliters (uL) in the "Soil
            Extract Volume" field on Form I VOA.   This volume includes any
            methanol not collected from the filtration of the extract through
            glass wool; the volume is typically 10,000 uL (i.e., the 10 mL of
            methanol used for the extraction).  If a medium soil sample is
            analyzed, enter the volume of the methanol extract added to the
            reagent water in the purge tube and analyzed in the "Soil Aliquot
            Volume" field.  Enter this volume in microliters (uL).

3.4.2.10    For semivolatiles and pesticides/Aroclors, enter the actual volume
            of the most concentrated sample extract, in microliters (uL),  in
            the "Concentrated Extract Volume" field on Form I SV-1, SV-2 or
            PEST.  For semivolatiles, this volume will typically be 1,000 uL
            (for water) or 500 uL (for water and soil) when GPC is performed.
            For pesticides/Aroclors, the volume of the most concentrated
            extract will typically be 10,000 uL (for water) or 5,000 uL (for
            water and soil) when GPC is performed.  For pesticides/Aroclors,
            the volume of the most concentrated extract is not the volume
            taken through the Florisil and sulfur cleanup steps.  If a
            dilution of the sample extract is made in a subsequent analysis,
            this volume will remain the same, but the dilution factor will
            change.

3.4.2.11    For semivolatiles and pesticides/Aroclors, enter the volume of the
            sample extract injected into the GC in the "Injection Volume"
            field on Form I SV-1, SV-2 or PEST.  Report this volume in
            microliters (uL) to one decimal place (e.g., 1.0 uL) .

            NOTE:  A 2.0 microliter injection is required for semivolatile
            analyses.

3.4.2.12    If pesticides/Aroclors are analyzed using two GC columns connected
            to a single injection port, enter the amount of half the volume  in
            the syringe in the "Injection Volume" field  (i.e., assume that the
            extract injected is evenly divided between the two columns).

3.4.2.13    If a sample or sample extract has been diluted for analysis, enter
            the dilution factor as a single number  (e.g., enter 100.0 for a  1
            to 100 dilution of the sample)  in the "Dilution Factor" field.
            The dilution factor shall not be entered  as a fraction.  If a
            sample was not diluted, enter 1.0.  Report dilution factors to one
            decimal place.

3.4.2.14    If sulfur cleanup is employed,  enter Y  in the  "Sulfur  Cleanup"
            field; if not, enter N on Form  I PEST.
                                     B-38                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                        Form I
3.4.2.15    For positively identified target compounds, the Contractor shall
            report the concentrations as uncorrected for blank contaminants.

3.4.2.16    For volatile and semivolatile results, report analytical results
            to one significant figure if the value is less than 10, and two
            significant figures if the value is 10 or above.  Report all
            pesticide/Aroclor results to two significant figures.

3.4.2.17    Enter the appropriate concentration units,  ug/L or ug/Kg.

3.4.2.18    Under the column labeled "Q" for qualifier, flag each result with
            the specific data reporting qualifiers listed below.  When
            reporting results to EPA, the Contractor shall use these contract-
            specific qualifiers.  The Contractor shall  not modify the
            qualifiers.  Up to five qualifiers may be reported on Form I for
            each compound.   The Contractor is encouraged to use additional
            flags or footnotes (see the X qualifier).

            The EPA-defined qualifiers to be used are:

             U:  This flag indicates the compound was analyzed for but not
                 detected.   The CRQL shall be adjusted  according to the
                 equation listed in Exhibit D.   CRQLs are listed in Exhibit C.

             J:  This flag indicates an estimated value   This flag is used
                 (1)  when estimating a concentration for tentatively
                 identified compounds where a 1:1 response is assumed, (2)
                 when the mass spectral and retention time data indicate the
                 presence of a compound that meets the  volatile and
                 semivolatile GC/MS identification criteria, and the result is
                 less than the CRQL but greater than zero,  and (3)  when the
                 retention time data indicate the presence of a compound that
                 meets the pesticide/Aroclor identification criteria, and the
                 result is less than the CRQL but greater than zero.  For
                 example,  if the sample quantitation limit is 10 ug/L, but a
                 concentration of 3 ug/L is calculated,  report it  as 3J.

                 NOTE:   The J flag is not used  and the  compound is  not
                 reported as being identified for pesticide/Aroclor results
                 less than the CRQL if the pesticide residue analysis expert
                 determines that the peaks used for compound identification
                 resulted from instrument noise or other interferences (column
                 bleed,  solvent contamination,  etc.).

             N:  This flag  indicates presumptive evidence of a compound.   This
                 flag is only used for tentatively identified compounds
                 (TICs),  where the identification is based on a mass spectral
                 library search.   It is applied to all  TIC results.   For
                 generic characterization of a  TIC,  such as chlorinated
                 hydrocarbon,  the N flag is not used.
                                     B-39                               OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form I
             P:  This flag is used for a pesticide/Aroclor  target analyte when
                 there is greater than 25% difference for detected
                 concentrations between the two GC columns  (see Form X).   The
                 lower of the two values is reported on Form I  and flagged
                 with a P.

             C:  This flag applies to pesticide results where the
                 identification has been confirmed by GC/MS.  If GC/MS
                 confirmation was attempted but was unsuccessful, do not  apply
                 this flag; use a laboratory-defined flag instead (see the X
                 qualifier).

             B:  This flag is used when the analyte is found in the associated
                 method blank as well as in the sample.   It indicates probable
                 blank contamination and warns the data user to take
                 appropriate action.  This flag shall be used for a
                 tentatively identified compound as well as for a positively
                 identified target compound.

                 The combination of flags BU or UB is expressly prohibited.
                 Blank contaminants are flagged B only when they are detected
                 in the sample.

             E:  This flag identifies compounds whose concentrations exceed
                 the upper level of the calibration range of the instrument
                 for that specific analysis.   If one or more compounds have a
                 response greater than the upper level of the calibration
                 range, the sample or extract shall be diluted  and reanalyzed
                 according to the specifications in Exhibit D;  exceptions are
                 also noted in Exhibit D.  All such compounds with a response
                 greater than the upper level of the calibration range shall
                 have the concentration flagged with an E on Form I for the
                 original analysis.

                 NOTE:  For total xylenes, where three isomers  are quantified
                 as two peaks, the calibration range of each peak shall be
                 considered separately.  For example, a diluted analysis  is
                 not required for total xylenes unless the  concentration  of
                 the peak representing the single isomer exceeds 200 ug/L or
                 the peak representing the two co-eluting isomers on that GC
                 column exceeds 400 ug/L.  Similarly, if the two 1,2-
                 Dichloroethene isomers coelute, a diluted  analysis is not
                 required unless the concentration exceeds  400  ug/L.

             D:  If a sample or extract is reanalyzed at a  higher dilution
                 factor, for example when the concentration of  an analyte
                 exceeds the upper calibration range, the DL suffix is
                 appended to the sample number on Form I for the more diluted
                 sample, and all reported concentrations on that Form I are
                 flagged with the D flag.  This flag alerts data users that
                 any discrepancies between the reported concentrations may be
                 due to dilution of the sample or extract.   NOTE 1:  The D
                                     B-40                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                        Form I
                  flag is not applied to compounds which are not detected in
                  the sample analysis i.e. compounds reported with the CRQL and
                  the U flag.  NOTE 2:  Separate Form Is are required for
                  reporting the original analysis (EPA Sample No. XXXXX) and
                  the more diluted sample analysis (EPA Sample No. XXXXXDL)
                  i.e. the results from both analyses cannot be combined on a
                  single Form I.

              A:   This flag indicates that a tentatively identified compound is
                  a suspected aldol-condensation product.

              X:   Other specific flags may be required to properly define the
                  results.  If used, the flags shall be fully described, with
                  the description attached to the sample data summary package
                  and the SDG Narrative.  Begin by using X.  If more than one
                  flag is required, use Y and Z as needed.  If more than five
                  qualifiers are required for a sample result, use the X flag
                  to represent a combination of several flags.  For instance,
                  the X flag might combine the A, B,  and D flags for some
                  samples.  The laboratory-defined flags are limited to X, Y,
                  and Z.

3.5   Organic Analysis Data Sheet:  Tentatively Identified Compounds (Form I
      VOA-TIC and Form I SV-TIC)

3.5.1    Purpose.  This form is used to report analysis results for non-target
         compounds (e.g., compounds not listed in Exhibit C), excluding system
         monitoring compounds,  surrogates and internal standards.  See Exhibit
         D for instructions on identification and quantitation.  The
         Contractor shall submit Form I VOA-TIC or SV-TIC for every analysis,
         including required dilutions and reanalyses, even if no TICs are
         found.

3.5.2    Instructions.   Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions in addition to the instructions in
         Section  3.4.

3.5.2.1     Report all  TICs including CAS number (if applicable), compound
            name, retention time,  and the estimated concentration as
            uncorrected for blank contaminants.   If the analytical result is
            less than 10, report to one significant figure.   If the analytical
            result is 10 or greater,  report to two significant figures.
            (Criteria for reporting TICs are given in Exhibit D, Section 11).
            Retention time shall be reported in minutes and decimal minutes,
            not seconds or minutes:seconds.   If, in the opinion of the mass
            spectral  interpretation specialist,  no valid tentative
            identification can  be made, the compound shall be reported as
            unknown.
                                     B-41                              OLM03.0

-------
 Exhibit  B  —  Section  3
 Forms  Instructions
 Form II
 3.5.2.2     Total the number of TICs found, including aldol-condensation
            products  (see Section 3.5.2.4), and enter this number in the
            "Number TICs found" field.  If no TICs were found, enter 0  (zero).

 3.5.2.3     If the name of a compound exceeds the 28 spaces in the TIC  column,
            truncate the name to 28 characters.  If the compound is an
            unknown, restrict the description to no more than 28 characters
            (e.g., unknown hydrocarbon).

 3.5.2.4     Peaks that are suspected to be aldol-condensation reaction
            products  (e.g., 4-methyl-4-hydroxy-2-pentanone and
            4-methyl-3-pentene-2-one) shall be summarized on this form,
            flagged A, and included in the "Number TICs found" field.   The
            peaks shall be counted as part of the 30 most intense non-target
            semivolatile compounds to be searched.

 3.6   System Monitoring Compound Recovery (Form II VOA)

 3.6.1    Purpose.  For volatiles,  Form II VOA is used to report the recoveries
         of the system monitoring compounds added to each volatile sample,
         including dilutions and reanalyses,  blank, matrix spike, and matrix
         spike duplicate.  The system monitoring compounds are used to monitor
         the performance of the purge and trap-gas chromatograph-mass
         spectrometer system as a whole.  Form II VOA is matrix-specific, so
         that system monitoring compound recoveries for \ fater samples are
         reported on a different version of Form .1 than the recoveries for
         soil samples.  Soil sample recoveries are further differentiated by
         concentration level.

3.6.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete one form for each level.  Do
         not mix low and medium level samples on one form.  NOTE:  For
         volatile soil samples only,  specify  the level as LOW or MED.
         Complete the remainder of the form using the following instructions.

3.6.2.1     For each system monitoring compound listed in Table 1,  report the
            percent recovery to the nearest whole percentage point,  and to the
            number of significant  figures given by the QC limits at the bottom
            of the form.

3.6.2.2     Flag each system monitoring compound recovery outside the QC
            limits with an asterisk (*).  The asterisk shall be placed  in the
            last space in each appropriate column, under the "#" symbol.

3.6.2.3     In the "TOT OUT" column,  total the number of system monitoring
            compound recoveries that were outside the QC limits for each
            sample.   If no system  monitoring  compounds were outside the
            limits,  enter 0 (zero).

3.6.2.4     Number all pages as described in  Section 3.3.
                                     B-42                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                       Form II
                                    Table 1
                          System Monitoring  Compounds
            Volatile System                     CAS Number
            Monitoring Compounds

            SMC  1: Toluene-d8  (TOL)             2037-26-5
            SMC  2: Bromofluorobenzene  (BFB)     460-00-4
            SMC  3: l,2-Dichloroethane-d4  (DCE)  17060-07-0
3.7   Surrogate Recovery (Form II SV and Form II PEST)

3.7.1    Purpose.  Form II SV and Form II PEST are used to report the
         recoveries of the surrogate compounds added to each semivolatile and
         pesticide/Aroclor sample, blank, matrix spike, and matrix spike
         duplicate.  Form II SV and Form II PEST are matrix-specific as well
         as fraction-specific, so surrogate recoveries for semivolatile and
         pesticide water samples are reported on a different version of Form
         II than surrogate recoveries for semivolatile and pesticide soil
         samples.

3.7.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  NOTE:  For semivolatile soil samples
         only, specify the level as LOW or MED.   Complete one form for each
         level.  Do not mix low and medium level samples on one form.
         Complete the remainder of the form using the following instructions.

3.7.2.1        For each surrogate listed in Tables 2 and 3, report the percent
               recovery to the nearest whole percentage point.

3.7.2.2        Flag each surrogate recovery outside the QC limits with an
               asterisk (*).  The asterisk shall be placed in the last space
               in each appropriate column,  under the "#" symbol.

3.7.2.3        In the "TOT OUT" column,  total the number of surrogate
               recoveries that were outside the  QC limits for each sample.  If
               no surrogates were outside the limits, enter 0 (zero).

3.7.2.4        If the sample is diluted and the  surrogates are outside the
               acceptance window in any analysis, enter the calculated
               recovery, and flag the surrogate  recoveries with a D in the
               column under the "#" symbol.  Do  not include results flagged
               with a D in the total number of recoveries for each sample
               outside the QC limits.

3.7.2.5        The pesticide surrogate recoveries shall be reported from both
               GC columns used for the analyses.  Therefore, identify each GC
               column at the top of Form II PEST, entering the stationary


                                     B-43                              OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form II
               phase in the "GC Column" field, and the internal diameter of
               the column in millimeters (mm) in the "ID" field.

3.7.2.6        The assignment of columns as "1" and "2" is left to the
               discretion of the Contractor when the analyses are performed by
               simultaneous injection into a GC containing two columns.  If so
               analyzed, the assignment of "GC Column 1" and "GC Column 2"
               shall be consistent across all the reporting forms.  If the
               analysis is not performed by simultaneous injection, then the
               assignment of GC column number shall be based on the
               chronological order of the two analyses.

3.7.2.7        Although the pesticide surrogate recovery limits for samples,
               matrix spike and matrix spike duplicates are only advisory, the
               Contractor shall flag those recoveries that are outside the
               advisory QC limits or are diluted out.  The total number of
               recoveries that are outside the QC limits shall include all
               values from both of GC columns.  In counting the total number
               of recoveries that are outside the QC limits, do not include
               the results flagged with a D.

3.7.2.8        Number all pages as described in Section 3.3.
                                   Table 2
                           Semivolatile Surrogates
         Semivolatile Surrogates              CAS Number


         SI: Nitrobenzene-d5  (NBZ)            4165-60-0
         S2: 2-Fluorobiphenyl  (FBP)           321-60-8
         S3: Terphenyl-dl4  (TPH)              98904-43-9
         S4: Phenol-d5  (PHL)                  4165-62-2
         S5: 2-Fluorophenol  (2FP)             367-12-4
         S6: 2,4,6-Tribromophenol  (TBP)       118-79-6
         S7: 2-Chlorophenol-d4  (2CP)          93951-73-6
         S8: l,2-Dichlorobenzene-d4  (DCB)     2199-69-1
                                   Table 3
                             Pesticide Surrogates
       Pesticide Surrogates                  CAS Number


       Decachlorobiphenyl  (DCB)              2051-24-3
       Tetrachloro-m-xylene  (TCX)            877-09-8



                                     B-44                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                      Form III
3.8   Matrix Spike/Matrix Spike Duplicate Recovery (Form III, All Fractions)

3.8.1    Purpose.  This form is used to report the results of the analyses of
         matrix spikes and matrix spike duplicates (MS/MSD).  The form is
         matrix-specific for volatiles, semivolatiles, and pesticides.

3.8.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Include the EPA sample number for the
         matrix spike, without the suffixes MS or MSD.  Complete the remainder
         of the form using the following instructions.

3.8.2.1     For volatile and semivolatile soil samples, specify level as LOW
            or MED on Form III VOA or SV.  SDGs containing soil samples at
            both levels require a MS/MSD at each level; therefore, for soils,
            prepare one form for each level.

3.8.2.2     In the first table under the "SPIKE ADDED" column, enter the
            calculated concentration in ug/L or ug/Kg  (according to the
            matrix) that results from dividing each spike compound amount
            added to the aliquot weight/volume chosen  for the matrix spike.
            For instance, for base/neutral compounds in medium level soils, if
            50 ug of spike are added to 1 g of soil, the resulting
            concentration is 50,000 ug/Kg.

3.8.2.3     Enter the sample concentration in the next column, in similar
            units,  of each spike compound detected in the original sample.  If
            a spike compound was not detected during the analysis of the
            original sample, enter the sample result as 0 (zero).

3.8.2.4     In the "MS CONCENTRATION" column, enter the actual concentration
            of each spike compound detected in the matrix spike aliquot.

3.8.2.5     Calculate the percent recovery of each spike compound in the
            matrix spike aliquot to the nearest whole percent, according to
            Exhibit D.  Enter the percent recovery in the "MS % REC" column.

3.8.2.6     Flag all percent recoveries outside the QC limits with an asterisk
            (*).  The asterisk shall be placed in the  last space of the "MS %
            REC" column, under the "#" symbol.

3.8.2.7     For pesticide/Aroclor matrix spikes and matrix spike duplicates,
            the MS concentration and MSD concentration shall be the
            concentration of the spiked analyte reported on Form I for those
            analyses.  Of the two concentrations calculated for each
            pesticide/Aroclor target compound, one on each GC column, the
            lower concentration shall be reported on Form I, and both
            concentrations shall be reported on Form X.  The lower
            concentration is also reported on Form III and used in the
            calculation of spike recovery, even if that concentration yields a
            recovery value that is outside the advisory QC limits.
                                     B-45                              OLM03.0

-------
 Exhibit B — Section 3
 Forms  Instructions
 Form IV
 3.8.2.8     Follow Sections 3.8.2.2 through 3.8.2.7 to complete the lower
            table, using the results of the analysis of the MSD aliquot.

 3.8.2.9     Calculate the relative percent difference (RPD) between the matrix
            spike recovery and the matrix spike duplicate recovery, and enter
            this value in the "% RPD" column.  Report the RPD to the nearest
            whole percent.

 3.8.2.10    Compare the RPDs to the QC limits given on the form, and flag each
            RPD outside the QC limits with an asterisk {*) in the last space
            of the "% RPD" column, under the "#" symbol.

 3.8.2.11    Summarize the values outside the QC limits at the bottom of the
            page.  No further action is required by the Contractor.

 3.9   Method Blank Summary (Form IV, All Fractions)

 3.9.1    Purpose.  This form summarizes the samples associated with each
         method blank analysis.   The Contractor shall submit the appropriate
         Form IV for each blank.

 3.9.2    Instructions.  Complete the header information according to the
         instructions in Section 3.3.   The EPA sample number entered in the
         upper righthand corner shall be the same number  entered on Form I for
         the blank.   Complete the remainder of the form using the following
         instructions.

 3.9.2.1     Complete the following fields:  "Instrument ID," "Date Analyzed,"
            and "Time Analyzed."  Dates shall be entered  as MM/DD/YY.   The
            time shall be reported in military time.

3.9.2.2     Pesticide/Aroclor contaminants shall meet the identification
            criteria requiring analysis of the blank on two different GC
            columns  (see Exhibit D PEST).   Enter the date, time and instrument
            ID of both analyses  of the blank on the pesticide method blank
            summary  (Form IV PEST).   The information on the two analyses is
            differentiated as Date Analyzed (1), Date Analyzed (2), etc.   If
            the analyses were run simultaneously,  the order of reporting is
            not important,  but shall be consistent with the information
            reported on all other pesticide forms.   Otherwise, Date Analyzed
            (1)  shall indicate the analysis on column 1,  and Date Analyzed (2)
            shall indicate the analysis on column 2.

3.9.2.3     Identify the GC column and internal diameter  in the appropriate
            fields.

3.9.2.4     For volatiles,  indicate the purging method by entering Y for
            heated purge or N for ambient  temperature purge in the "Heated
            Purge:  Y/N" field on Form IV VOA.
                                     B-46                              OLM03.1

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                        Form V
3.9.2.5     For pesticide/Aroclor blanks, enter the method of extraction as
            SEPF for separator/ funnel, SONC for sonication, or CONT for
            continuous liquid-liquid extraction on Form IV PEST.

3.9.2.6     For semivolatile and pesticide/Aroclor method blanks, enter the
            date of extraction of the blank on Form IV SV or PEST.

3.9.2.7     If the samples associated with pesticide/Aroclor blank are
            subjected to sulfur cleanup, then the blank shall also be
            subjected to sulfur cleanup.  If sulfur cleanup is employed, enter
            Y in the "Sulfur Cleanup" field; if not, enter N on Form IV PEST.
            If only some of the samples associated with the method blank are
            subjected to sulfur cleanup, a separate sulfur cleanup blank is
            required (see Exhibit D PEST).  If a separate sulfur cleanup blank
            is prepared, complete one version of Form IV associating all the
            samples with the method blank, and a second version of Form IV
            listing only  those samples associated with the separate sulfur
            cleanup blank.  NOTE:  Subjecting all samples associated with a
            method blank to sulfur cleanup avoids the need for two forms.

3.9.2.8     For all three fractions, as appropriate, summarize the samples
            including storage and volatile instrument blanks,  associated with
            a given method blank in the table, entering the EPA sample number
            and lab sample identifier.  For volatiles,  enter the lab file
            identifier and the time of analysis of each sample.  For
            semivolatiles, enter lab file identifier and the date of analysis.
            For pesticides/Aroclors, enter the dates of both analyses as Date
            Analyzed (1) and Date Analyzed (2), as discussed previously.

3.9.2.9     For pesticide/Aroclor fraction, enter the lab file identifier only
            if GC/MS confirmation was attempted.  Otherwise, leave this field
            blank.

3.9.2.10    Number all pages as described in Section 3.3.
3.10  GC/MS Instrument Performance Check and Mass Calibration (Form V VOA and
      Form V SV)

3.10.1   Purpose.  This form is used to report the results of the GC/MS
         instrument performance check for the volatile and semivolatile
         fractions and to summarize the date and time of analyses of samples,
         including dilutions and reanalyses, standards, blanks,  matrix spikes,
         and matrix spike duplicates associated with each analysis of the
         instrument performance check solution.

3.10.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.
                                     B-47                              OLM03.1

-------
Exhibit B — Section 3
Forms Instructions
Form VI
3.10.2.1    Enter the date and time of injection of the instrument performance
            check solution (BFB for volatiles—CAS Number 460004, DFTPP for
            semivolatiles—CAS Number 5074715).  The date shall be entered as
            MM/DD/YY.  The time shall be reported as military time.

3.10.2.2    For volatiles, identify the GC column and internal diameter on
            Form V VOA.

3.10.2.3    For volatiles, indicate the purging method by entering Y for
            heated purge or N for ambient temperature purge in the "Heated
            Purge: Y/N" field on Form V VOA.  Water samples and medium soil
            sample extracts may be reported on the same Form V if analyzed
            together, since a single calibration may be applied to both sample
            types.

3.10.2.4    For each ion listed on the form, enter the percent relative
            abundance in the righthand column of the first table.  Report
            relative abundances to the number of significant figures given for
            each ion in the ion abundance criteria column.

            NOTE:  For both BFB and DFTPP, one or more of the high mass ions
            may exceed the abundance of the ion listed on the form as the
            nominal base peak, m/z 95 for BFB and m/z 198 for DFTPP.  Despite
            this possibility,  all ion abundances shall be normalized to the
            nominal base peaks listed on Form V (see Exhibits D and E).

3.10.2.5    All relative abundances shall be reported as a number.  If the
            relative abundance is zero, enter 0, not a dash or other
            non-numeric character.  Where parentheses appear, compute the
            percentage of the ion abundance of the mass given in the
            appropriate footnote, and enter that value in the parentheses.

3.10.2.6    In the lower table, list all samples,  including dilutions and
            reanalyses, standards, blanks, matrix spikes, and matrix spike
            duplicates analyzed under that instrument performance check in
            chronological order,  by time of analysis (in military time).
            Refer to Section 3.3.7 for specific instructions for identifying
            standards and blanks.

3.10.2.7    Complete the following fields for all standards, samples,
            including dilutions and reanalyses, blanks, matrix spikes, and
            matrix spike duplicates:  "EPA Sample No.," "Lab Sample ID," "Lab
            File ID," "Date Analyzed," and "Time Analyzed."

3.10.2.8    Number all pages as described in Section 3.3.
                                     B-48                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                       Form VI
3.11  GC/MS Initial Calibration Data (Form VI VOA and Form VI SV)

3.11.1   Purpose.  After a GC/MS system has undergone an initial five-point
         calibration at the specific concentration levels described in Exhibit
         D, and after all initial calibration criteria have been met, the
         Contractor shall complete and submit this form for each volatile or
         semivolatile target compound initial calibration performed which is
         relevant to the samples, including dilutions and reanalyses, blanks,
         matrix spikes, or matrix spike duplicates in the SDG, regardless of
         when that calibration was performed.

3.11.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Enter the Case number and SDG number
         for the current data package,  regardless of the original Case for
         which the initial calibration was performed.  Complete the remainder
         of the form using the following instructions.

3.11.2.1    Enter the date(s) of the calibration.  If the calendar date
            changes during the calibration procedure, the inclusive dates
            shall be recorded.  Dates shall be entered as MM/DD/YY.

3.11.2.2    Enter the injection times of the first and last of the standards
            analyzed in the "Calibration Times" field.  Times shall be
            reported in military time.

3.11.2.3    For volatiles, complete the "Heated Purge," "GC Column," and "ID"
            fields as on Form V.

3.11.2.4    Enter the lab file identifier for each of the five calibration
            standards injected.  Complete the response factor data for the
            five calibration points, and then calculate and report the average
            relative response factor (RRF) for all target compounds.

3.11.2.5    For volatiles, report the relative response factors for the system
            monitoring compounds in the calibration standards.  For
            semivolatiles, report the response factors for all surrogate
            compounds in the calibration standards.  The Contractor shall
            report the relative standard deviation (%RSD) for all compounds.
            See Exhibit D for equations.
     1For semivolatiles,  eight compounds (2,4-Dinitrophenol,  2,4,5-
Trichlorophenol, 2-Nitroaniline, 3-Nitroaniline, 4-Nitroaniline,
4-Nitrophenol, 4,6-Dinitro-2-Methylphenol, and Pentachlorophenol)  will only
require a four-point initial calibration at 50, 80, 120, and 160 total
nanograms because detection at less than 50 ng per injection is difficult.  If
a four-point calibration is performed for these compounds, leave the "RRF20"
column blank.

                                     B-49                              OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form VI
3.12  GC/EC Initial Calibration Data (Form VI PEST)

3.12.1   Purpose.  The initial calibration of pesticides/Aroclors involves the
         determination of retention times, retention time windows, and
         calibration factors.  For single component pesticide target
         compounds, these data are calculated from the analyses of the
         Individual Standard Mixtures A and B at three different concentration
         levels.  For the multicomponent target compounds, these data are
         calculated from a single point calibration.

3.12.2   Instructions.  Complete one Form VI for each GC column used for the
         three analyses of Individual Standard Mixture A (low point, mid-
         point, and high point) and the three analyses of Individual Standard
         Mixture B during an initial calibration.  Complete the header
         information according to the instructions in Section 3.3.  Complete
         the remainder of the form using the following instructions.

3.12.2.1    In the "Level (x low)" field,  enter the concentration of the low
            point, mid-point, and high point calibration standards as a
            multiplier of the low point.  Therefore, for the low point, enter
            "1.0."  The concentration of the mid-point standard is specified
            in Exhibit D as four times the low point; therefore, enter "4.0."
            The high point standard shall  be at least 16 times the low point,
            but may be higher,  if that value lies within the linear range of
            the instrument,  as specified in Exhibit D.  Therefore, enter the
            appropriate multiplier to the  high point standard concentration to
            one decimal place.

3.12.2.2    Identify the GC column and internal diameter (in millimeters, mm)
            in the appropriate fields.

3.12.2.3    Enter the dates of analysis of the first and last of the six
            standards on each form in the  "Date(s) Analyzed" field.  Dates
            shall be entered as MM/DD/YY.

3.12.2.4    For each standard analyzed, enter the retention time of each
            applicable analyte in minutes  and decimal minutes, under the
            appropriate concentration level in the "RT OF STANDARDS" column on
            Form VI PEST-1.

3.12.2.5    Calculate the mean retention time of each analyte from the three
            individual mixtures,  and report it in the "MEAN RT" column on Form
            VI PEST-1.

3.12.2.6    Calculate the retention time window for each analyte using the
            specifications in Exhibit D, and enter the lower limit of the
            window in the "RT WINDOW" column under "FROM," and the upper limit
            of the window under "TO" on Form VI PEST-1.  The retention times
            of the surrogates are reported from the analyses of Individual
            Mixture A and the windows are  only required to be calculated for
            Individual Mixture A.
                                     B-50                              OLM03.0

-------
                                                        Exhibit  B  —  Section 3
                                                            Forms  Instructions
                                                                        Form VI
 3.12.2.7     For the  six  analyses  of the  Individual Standard Mixtures, the
             Contractor shall  also complete the calibration factor data on Form
             VI PEST-2.   Prepare one form for each instrument and GC column
             used.  Enter the  calibration factor for each compound in each of
             the standards.  Calculate  and enter a mean calibration factor and
             a relative standard deviation (%RSD).  As with surrogate retention
             times, the surrogate  calibration factors are only required from
             Individual Mixture A  analyses.

 3.12.2.8     For the  multicomponent target compounds, the retention times,
             retention time windows, and  calibration factors shall be reported
             in a similar fashion  for each single point calibration standard.
             For each multicomponent compound, the Contractor shall select at
             least three  peaks from each  analyte, according to the
             specifications in Exhibit  D.  The retention time and calibration
             factor data  apply to  each  peak.  Complete one version of Form VI
             PEST-3 for each GC column, for each initial calibration that
             applies  to samples in the  data package.

 3.12.3   Form VI is  also used to  report  the results of analysis of the
         Resolution  Check Solution that  shall begin each pe?ticide/Aroclor
         initial calibration sequence  (Form VI PEST-4).  The Contractor shall
         submit one  Form VI PEST-4 for both GC columns.

 3.12.4   Complete the header information  as described in Section 3.3.  Using
         the same assignment of first  and second GC columns made for Form IV,
         enter the GC column identifier,  internal diameter, and date and time
         of analysis.  Enter the EPA sample number for the Resolution Check
         Standard.    If simultaneous injections on a single GC are used,  the
         EPA sample  number may be the  same for both Resolution Check
         Standards.   If  simultaneous injections are not used,  use different
         suffixes to identify the standards.   Complete the remainder of the
         form using the  following instructions.

3.12.4.1       List each  analyte,  in retention time order,  including both
               surrogate  compounds.  Thus, the order of analytes in the two
               boxes on  this form will be different due to the dissimilarity
               of the stationary phases of the two GC columns used.   Enter the
               name of each target analyte in the Resolution Check Mixture as
               it appears on Form I PEST.  Spell out the names of the
               surrogates as they appear on Form VII PEST-2.

3.12.4.2       Enter the retention time of each analyte from the analysis in
               the  "RT"   column.

3.12.4.3       Calculate the resolution between each pair of analytes.   Enter
               the  resolution between the first and second peaks on the  line
               for  the first analyte listed in the box.   Enter the resolution
               between the second  and third peaks on the line for the second
               analyte,  and so on, until the  resolutions of all possible pairs
               of adjacent analytes have been entered.   NOTE:   Only eight of
               the  nine  resolution fields will be filled.
                                     B-51                              OLM03.0

-------
Exhibit  B  — Section 3
Forms  Instructions
Form VII
3.12.4.4       Form VI  (PEST-5, PEST-6 and PEST-7 for each pair of PEM, mid-
               level initial calibration mixture A, and mid-level initial
               calibration mixture B, respectively) shall be used to report
               the percent resolution between each pair of analytes according
               to the definition in Exhibit D (Pesticides).

3.12.4.5       Complete the header information as described in Section 3.3.
               Using the same assignment of first and second GC columns made
               for Form IV, enter the GC column identifier, internal diameter,
               and date and time of analysis.  Enter the EPA sample number for
               the respective standards.  If simultaneous injections are not
               used, use different suffixes to identify the standards.
               Complete the remainder of the form using the following
               instructions.

3.12.4.5.1     List each analyte,  in retention time order, including both
               surrogate compounds.  Thus, the order of analytes in the two
               boxes on this form will be different due to the dissimilarity
               of the stationary phases of the two GC columns used.   Enter the
               name of each target analyte in the standard as it appears on
               Form I PEST.  Spell out the names of the surrogates as they
               appear on Form VII  PEST-2.

3.12.4.5.2     Enter the retention time of each analyte from the analysis in
               the "RT" column.

3.12.4.5.3     Calculate the resolution between each pair of analytes.  Enter
               the resolution between the first and second peaks on the line
               for the first analyte listed in the box.   Enter the resolution
               between the second  and third peaks on the line for the second
               analyte, and so on,  until the resolutions of all possible pairs
               of adjacent analytes have been entered.   NOTE: The last
               resolution field will be left blank in each table.

3.13  GC/MS Continuing Calibration Data (Form VII VGA and Form VII SV)

3.13.1   Purpose.  For volatiles and semivolatiles, this form is used to
         report the calibration of the GC/MS system by the analys_s of
         specific calibration standards.  Form VII is required for each 12-
         hour time period for both volatile and semivolatile target compound
         analyses.  The Contractor shall analyze calibration standards and
         meet all criteria outlined in Exhibit D for the minimum RRF and
         maximum percent difference between initial and continuing
         calibrations.

3.13.2   Instructions.   Complete the header information according to the
         instructions in Section 3.3.   Complete the remainder of the form
         using the following instructions.

3.13.2.1    Enter the date and time of the continuing calibration and the
            dates and times of the initial calibration (give inclusive dates
            if the initial calibration is performed over more than one date).
                                     B-52                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                      Form VII
            Dates shall be entered as MM/DD/YY.  Times shall be reported in
            military time.

3.13.2.2    For volatiles, enter the pu*-ge method, GC column identifier, and
            internal diameter as on Form V.

3.13.2.3    Using the appropriate initial calibration (volatile or
            semivolatile), enter the average relative response factor (RRF)
            for each target compound, for each system monitoring compound for
            volatiles, and for each surrogate for semivolatiles.

3.13.2.4    Report the relative response factor (RRF50)  from the continuing
            calibration standard analysis.

3.13.2.5    Calculate the percent difference (%D)  for all compounds.  See
            Exhibit D for equation.  If the %D is greater than 999.9, report
            as 999.9.  If the %D is less than -99.9, report as -99.9.

3.14  GC/EC Calibration Verification Summary (Form VII,  PEST-1, PEST-2)

3.14.1   Purpose.  Form VII is used to report the results of the Performance
         Evaluation Mixtures (PEMs) and the mid-point concentrations of
         Individual Standard Mixtures A and B  that, along with the PEM,
         bracket each 12-hour period of sample analyses.   The Contractor shall
         submit Form VII PEST-1 for each 12-hour sequence analyzed.  Form VII
         PEST-2 shall be completed each time the Individual Standard Mixtures
         are analyzed, for each GC column used.

3.14.2   Instructions.  Complete Form VII PEST-1 and PEST-2 for each standard
         reported on Form VIII PEST.  Complete the header information
         according to the instructions in Section 3.3.  Complete the remainder
         of the form using the following instructions.

         FORM VII PEST-1

3.14.2.1    Enter the date(s) of the initial calibration(s).  Give inclusive
            dates if the initial calibration is performed over more than one
            day.  Dates shall be entered as MM/DD/YY.

3.14.2.2    Identify the GC column and internal diameter in the appropriate
            fields.

3.14.2.3    On Form VII PEST-1,  enter the EPA sample number, lab sample
            identifier and date and time of analysis for the instrument blank
            that preceded the 12-hour sequence (PIBLK).   For the PEM that
            initiated or terminated the 12-hour sequence (PEM), enter the EPA
            sample number, lab sample identifier,  and date and time of
            analysis.

3.14.2.4    When reporting data for the PEM at the beginning of the initial
            calibration sequence,  leave the "EPA Sample No.," "Lab Sample ID,"
            "Date Analyzed," and "Time Analyzed" fields blank for the
                                     B-53                              OLM03.0

-------
 Exhibit  B  —  Section  3
 Forms  Instructions
 Form VII
             instrument blank  (PIBLK), when no instrument blank is analyzed
             before the PEM.  When reporting all other PEM analyses, the
             instrument blank fields shall be completed.

 3.14.2.5     In the table, report the retention time for each analyte in the
             PEM as well as the retention time windows.

 3.14.2.6     For each analyte in the PEM, enter the amount of the analyte found
             in the PEM, in nanograms (ng) to three decimal places, in the
             "CALC AMOUNT" column.

 3.14.2.7     Enter the nominal amount of each analyte in the PEM in the "NOM
             AMOUNT" column.

 3.14.2.8     Calculate the percent difference between the calculated amount and
             nominal amount for each analyte according to Exhibit D.  Report
             the values in the "%D" column.  If the %D is greater than 999.9,
             report as 999.9.  If the %D is less than -99.9,  report as -99.9.

 3.14.2.9     Calculate the percent breakdown for endrin and 4,4'-DDT and the
             combined percent breakdown in the PEM according to Exhibit D.
             Enter the values for the breakdown of endrin and 4,4'-DDT in their
             respective fields immediately under the table.

             FORM VII PEST-2

 3.14.2.10    The upper table on Form VII PEST-2 contains the retention time and
             amount data for Individual Standard Mixture A compounds.   The
             lower table contains the data for Mixture B.  Complete the form
             using the instructions in Sections 3.14.2.1 through 3.14.2.8 for
             Form VII PEST-1.

 3.15  Internal Standard Area and RT Summary (Form VIII VOA and Form VIII SV)

 3.15.1   Purpose.   This form is used to summarize the peak areas and retention
         times of the internal standards added to all volatile and
         semivolatile samples,  including dilutions and reanalyses, blanks,
         matrix spikes,  and matrix spike duplicates.   The data are used to
         determine when changes in internal standard responses will adversely
         affect quantification of target compounds.   This form shall be
         completed each time a continuing calibration is performed, or when
         samples are analyzed under the same GC/MS instrument performance
         check as an initial calibration.

3.15.2   Instructions.   Complete the header information according to Section
         3.3.  Complete the remainder of the form using the following
         instructions.   If samples are analyzed immediately following an
         initial calibration,  before another instrument performance check and
         a continuing calibration,  Form VIII shall be completed on the basis
         of the internal  standard areas of the 50 ug/L initial calibration
         standard for volatiles,  and the 50 ng initial calibration standard
         for semivolatiles.   Use the date and time of analysis of this
                                     B-54                              OLM03.0

-------
                                                         Exhibit B — Section 3
                                                             Forms Instructions
                                                                      Form VIII
          standard and the lab file identifier and areas  in  place of those of a
          continuing calibration standard.

 3.15.2.1     Enter the date and time of analysis  of the continuing calibration
             standard.   The date shall  be entered as MM/DD/YY.   The time shall
             be  reported as military time.

 3.15.2.2     For volatiles,  enter the purge  method,  GC column  identifier,  and
             internal  diameter as on Form V.

 3.15.2.3     From the  results  of the analysis  of  the continuing  calibration
             standard,  enter the area measured for each internal  standard  and
             its retention  time (in decimal  minutes)  under the appropriate
             column  in  the  "12  HOUR STD"  row.

 3.15.2.4     For each  internal  standard listed in  Tables 4 and 5,  calculate the
             upper limit  of  the area as the  area  of  the particular  standard
             plus  100 percent  of its area (i.e.,  two times the area in the "12
             HOUR  STD"  field),  and  the  lower limit  of the area as the area of
             the internal standard  minus  50  percent  of its area  (i.e., one half
             the area in  the "12  HOUR STD" field).   Report these values  in the
             "UPPER LIMIT"  and  "LOWER LIMIT" rows,  respectively.  Calculate the
             upper limit  of  the  retention time  as  the retention of  the internal
             standard plus  0.50  minutes  (30  seconds), and the lower limit  of
             the retention time  as  the  retention time in ;he standard minus
             0.50 minutes (30 seconds).

3.15.2.5     For each sample, including dilutions, reanalyses, blanks, matrix
             spikes, and matrix  spike duplicates, analyzed under a  given
             continuing calibration,  enter the EPA sample number and the area
            measured for each  internal standard and its retention  time.   If
            the internal standard  area is outside the upper or lower limits
            calculated in step 4,  flag that area with an asterisk  (*).  The
            asterisk shall be placed in the far righthand space of the box for
            each internal standard  area, directly under the "#" symbol.
            Similarly, flag the retention time of any internal standard that
            is outside the limits with an asterisk.

3.15.2.6    Number all pages as described in Section 3.3.

                                    Table 4
                         Volatile  Internal  Standards


        Volatile Internal Standards       CAS Number

        IS1:  Bromochloromethane (BCM)     74-97-5

        IS2:  1,4-Difluorobenzene (DFB)     540-36-3

        IS3:  Chlorobenzene-d5  (CBZ)       3114-55-4
                                     B-55                              OLM03.0

-------
Exhibit  B  —  Section  3
Forms  Instructions
Form VIII
                                    Table 5
                        Semivolatile Internal Standards
          Semivolatile Internal Standards            CAS Number


        IS1:  l,4-Dichlorobenzene-d4  (DCB)    3855-82-1
        IS2:  Naphthalene-d8  (NPT)            1146-65-2
        IS3:  Acenaphthene-dlO  (ANT)          15067-26-2
        IS4:  Phenanthrene-dlO  (PHN)          1517-22-2
        IS5:  Chrysene-dl2  (CRY)              1719-03-5
        IS6:  Perylene-dl2  (PRY)              1520-96-3


3.16  Pesticide Analytical Sequence  (Form VIII PEST)

3.16.1   Purpose.  This form is used to report the analytical sequence for
         pesticide analysis.  At least one form is required for each GC column
         used for pesticide/Aroclor analyses.

3.16.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.

3.16.2.1    Enter the date(s) of the initial calibration.  Give inclusive
            dates if the initial calibration is performed over more than one
            day.  Dates shall be entered as MM/DD/YY.

3.16.2.2    Identify the GC column and internal diameter in the appropriate
            fields.

3.16.2.3    At the top of the table, report the mean retention time for
            tetrachloro-m-xylene (TCX) and decachlorobiphenyl (DCB) calculated
            from the initial calibration sequence.

3.16.2.4    For every analysis associated with a particular analytical
            sequence starting with the initial calibration, enter the EPA
            sample number,  lab sample identifier,  and date and time of
            analysis.  Each sample analyzed as part of the sequence shall be
            reported on Form VIII PEST even if it is not associated with the
            SDG.  The Contractor shall use ZZZZZ as the EPA sample number to
            distinguish all samples that are not part of the SDG being
            reported.

3.16.2.5    Report the retention time of the surrogates for each analysis in
            the "TCX RT" and "DCB RT" columns.   All sample analyses shall be
            bracketed by acceptable analyses of instrument blanks, a PEM, and
            Individual Standard Mixtures A and B.   Given the fact that the
            initial  calibration may remain valid for some time (see Exhibit
            D),  it is only necessary to report the data from 12-hour periods

                                     B-56                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                     Form VIII
            when samples, dilutions, reanalyses, matrix spike, matrix spike
            duplicate, blanks, or multicomponent analytes for the 72 hour
            confirmation requirement in an SDG were analyzed.  The Contractor
            shall submit Form VIII for the initial calibration sequence and
            forms that include the PEMs and Individual Standard Mixtures that
            bracket any and all samples in the SDG.  While the data for time
            periods between the initial calibration and samples in the SDG are
            not a routine deliverable, the data shall be available as
            requested (e.g., at on-site evaluations).  Non-EPA samples shall
            be numbered ZZZZZ.

3.16.2.6    Flag all those values which do not meet the contract requirements
            by entering an asterisk (*) in the "RT" column, under the "#"
            symbol.  If the retention time cannot be calculated due to
            interfering peaks, leave the "RT" column blank for that surrogate,
            enter an asterisk in the last column, and document the problem in
            the SDG Narrative.

3.16.2.7    If more than a single copy of Form VIII PEST is required, enter
            the same header information on all subsequent pages for that GC
            column and instrument, and number each page as described in
            Section 3.3.
                                     B-57                              OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form IX
3.17  Pesticide Cleanup Summary (Form IX, PEST-1,  PEST-2)

3.17.1   Purpose.  This form summarizes the results of the checks performed
         for both cleanup procedures employed during the preparation of
         pesticide extracts for analysis.  Form IX PEST-1 is used to report
         the results of the check of the Florisil  cartridges used to process
         all sample extracts  and to associate the lot of cartridges with
         particular sample results so that problems with a particular
         cartridge lot may be tracked across all associated samples.  Form IX
         PEST-2 summarizes the results of the calibration of the Gel
         Permeation Chromatography (GPC) device that shall be used to process
         all soil sample extracts for pesticide/Aroclor analyses.

3.17.2   Instructions.  Complete the header information according to the
         instructions in Section 3.3.  Enter the Case number and SDG number
         for the current data package, regardless  of the original Case for
         which the cartridge check was performed.   Complete the remainder of
         the form using the following instructions.

            FORM IX PEST-1

3.17.2.1    Enter the Florisil cartridge lot number.

3.17.2.2    Enter the date the Florisil cartridge  check solution was analyzed
            in the "Date of Analysis" field.  The  date shall be entered as
            MM/DD/YY.

3.17.2.3    Complete the "GC Column" and "ID" fields for the two GC columns
            used to analyze the samples, including blanks, matrix spikes, and
            matrix spike duplicates.  Report all results from either GC column
            1 or GC column 2.

3.17.2.4    In the first table, enter the amount of spike added and spike
            recovered in nanograms (ng) for each analyte.

3.17.2.5    Calculate the percent recovery to the  nearest whole percent, and
            enter the number in the "% REC" field.  Flag each spike recovery
            outside the QC limits (shown on the form) with an asterisk  (*).
            The asterisk shall be placed in the last space in the "% REC"
            column, under the "#" symbol.

3.17.2.6    In the second table, complete the "EPA Sample No.," the "Lab
            Sample ID," and "Date Analyzed" fields for each sample and blank
            that were cleaned up using this lot of Florisil cartridges.

3.17.2.7    Number the pages as described in Section 3.3.

            FORM IX FSST-2

3.17.2.8    On Form IX PEST-2, enter an identifier for the GPC column and the
            date of calibration in the appropriate fields.
                                     B-58                              OLM03.0

-------
                                                        Exhibit B — Section 3
                                                            Forms Instructions
                                                                        Form X
3.17.2.9    Complete the "GC Column" and "ID" fields as on Form IX PEST-1 for
            florisil.  Report all results from a single column.

3.17.2.1C   For each of the pesticide matrix spike compounds listed in the
            first table, enter the amount of the spike added to the GPC column
            and the amount recovered, in nanograms (ng).

3.17.2.11   Calculate the percent recovery of each analyte, and enter these
            values on the form, to the nearest percent.  Compare the
            recoveries to the QC limits shown on the form, and flag all those
            values outside the limits with an asterisk (*) in the "% REC"
            column under the "#" symbol.

3.17.2.12   For each sample in the data package that was subjected to GPC
            under this calibration, enter the EPA sample number, lab sample
            identifier, and the date of both analyses in the second table.

3.17.2.13   If more than one copy of Form IX PEST-2 is required, number all
            pages as described in Section 3.3.

3.18  Pesticide/Aroclor Identification (Form X,  PEST-1, PEST-2)

3.18.1   Purpose.  This form summarizes the guantitations of all target
         pesticides/Aroclors detected in a given sample.  It reports the
         retention times of the compound on both columns on which it was
         analyzed, as well as the retention time windows of the standard for
         that compound on both of these columns.  In addition,  it is used to
         report the concentration determined from each GC column, and the
         percent difference between the two quantitative results.  Separate
         forms are used for single component analytes and multicomponent
         analytes.

         Form X is required for each sample, including dilutions and
         reanalyses, blank, matrix spike,  and matrix spike duplicate in which
         compounds listed in Exhibit C (Pesticides/Aroclors) are reported on
         Form I.   Do not generate a Form X for pesticide instrument blanks.

3.18.2   Instructions.   Complete the header information according to the
         instructions in Section 3.3.  Complete the remainder of the form
         using the following instructions.

3.18.2.1    Enter the date(s)  of analysis.   Dates shall be entered as
            MM/DD/YY.

3.18.2.2    Enter the GC column and internal diameter for each of the two
            columns.

3.18.2.3    For each single component pesticide positively identified, enter
            the name of the compound in the "ANALYTE" column as it appears on
            Form I.
                                     B-59                              OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form X
3.18.2.4    Enter the retention times on each column of the compounds detected
            in the sample next to the appropriate column designation  (1 or 2).

3.18.2.5    Enter the retention time windows on each column from the  initial
            calibration standard.  These data shall correspond with those on
            Form VI and shall be entered in a similar manner.  The lower value
            is entered under the "FROM" column, the upper value under the "TO"
            column.

3.18.2.6    Enter the concentration calculated from each GC column under the
            "CONCENTRATION" column.  Although the units are the same as those
            used on Form I, ug/L for water samples and ug/Kg for soil samples,
            do not enter any units on Form X.

3.18.2.7    Calculate the percent difference between the concentrations
            entered on this form.  See Exhibit D for equation, and report to a
            tenth of a percent in the "%D" column.  If the %D is greater than
            999.9, report it as 999.9.

3.18.2.8    The lower of the two concentrations is reported on Form I for each
            pesticide compound.  The lower concentration js used because, if
            present,  coeluting interferences are likely to increase the
            calculated concentration of any target compound.  If the percent
            difference between the calculated concentrations is greater than
            25.0 percent,  flag the concentration on Form I, as described
            previously.   This will alert the data user to the potential
            problems in guantitating this analyte.

3.18.2.9    If more pesticide compounds are identified in an individual sample
            than can be reported on one Form X, complete as many additional
            copies of Form X as necessary, duplicating all header information
            and numbering the pages as described in Section 3.3.

3.18.2.10   Report multicomponent analytes detected in samples on Form X PEST-
            2.  Complete the header information and GC column fields as
            described above.  For multicomponent analytes, it is necessary to
            report the retention time and concentration of each peak chosen
            for guantitation in the target analyte in a fashion similar to
            that for single component pesticides.  The concentrations of all
            peaks guantitated (three are reguired, up to five may be used) are
            averaged to determine the mean concentration.  Report the lower of
            the two mean concentrations on Form I.  Flag this value if the
            mean concentrations from the two GC columns differ by more than 25
            percent,  as described previously.

3.18.2.11   If more multicomponent compounds are identified in an individual
            sample than can be reported on one Form X, complete as many
            additional copies of Form X as necessary, duplicating all header
            information and numbering the pages as described in Section 3.3.
                                     B-60                              OLM03.0

-------
                                                         Exhibit  B —  Section 3
                                                             Forms Instructions
                                                                         Form X
 3.19   Sample Log-In Sheet (Form DC-1)

 3.19.1   Purpose.   This  form is used to document  the  receipt  and  inspection of
          sample  containers  and  samples.   One  original of  Form DC-1  is  required
          for each  sample shipping  container  (only the hardcopy form is
          required).   If  the samples  in  a  single  sample shipping  container  are
          assigned  to more than  one SDG,  the original  Form DC-1 shall be  placed
          with the  deliverables  for the  SDG of the lowest  alpha-numeric number,
          and a copy of Form DC-1 shall  be placed  with the deliverables for  the
          other SDGs.  The copies shall  be identified  as "copy(ies)," and the
          location  of the original  shall  be noted  on the copies.

 3.19.2   Instructions

 3.19.2.1     Sign and date the airbill  (if present).

 3.19.2.2     Complete the header information on the form,  including  the log-in
             date.

 3.19.2.3     Examine  the  shipping container and record the presence/absence  of
             custody  seals and their  condition (e.g.,  intact,  broken) in  item
             1.

 3.19.2.4     Record the custody  seal  numbers in item 2.

 3.19.2.5     Open the  container,  remove the enclosed sample documentation, and
             record the presence/absence  of chain-of-custody record(s), SMO
             forms  (e.g.,  Traffic Reports,  Packing Lists), and  airbills or
             airbill  stickers in  items 3-5.  Specify if there  is an  airbill
             present  or an airbill  sticker  in  item 5.   Record the  airbill or
             sticker  number  in item 6.

 3.19.2.6     Remove the samples  from  the  shipping container(s), examine the
             samples  and  the  sample tags  (if present),  and record  the condition
             of the sample bottles  (e.g.,  intact, broken,  leaking) and  presence
             of absence of sample tags in  items 7 and  8.

 3.19.2.7     Review the sample shipping documents and  compare the  information
             recorded on  all the documents  and  samples  and circle  the
             appropriate  answer  in  item 9.

 3.19.2.8     Record the date and time of  cooler receipt at the  laboratory in
             items 10 and  11.

 3.19.2.9     If there are  no problems observed  during  receipt,   sign  and date
             (include the  time) Form DC-1,  the  chain-of-custody record,  and the
            Traffic Report,  and write the  sample numbers on Form  DC-1  in the
             "EPA Sample #" column.

3.19.2.10   Record the appropriate sample tags and assigned laboratory
            numbers, if applicable.

3.19.2.11   Any comments  should be made in the "Remarks" column.

                                     B-61                              OLM03.0

-------
Exhibit B — Section 3
Forms Instructions
Form DC-2
3.19.2.12   Record the fraction designation (if appropriate) and the specific
            area designation (e.g., refrigerator number) in the "Sample
            Transfer" block.  Sign and date the "Sample Transfer" block.

3.19.2.13   Cross out unused columns and spaces.

3.19.2.14   If there are problems observed during receipt or an answer marked
            with an asterisk (e.g., "absent*") was circled, contact SMO and
            document the contact as well as resolution of the problem on a CLP
            Communication Log.   Following resolution, sign and date the forms
            and note, where appropriate, the resolution of the problem.

3.20  Document Inventory Sheet (Form DC-2)

3.20.1   Purpose.  The Document Inventory Sheet (Form DC-2) is used to record
         both the inventory of Complete SDG File (CSF) documents and the
         number of documents in the original sample data package which is sent
         to the EPA Region.

3.20.2   Instructions

3.20.2.1    Organize all EPA CSF documents as described in Exhibit B, Sections
            II and III.   Assemble the documents in the order specified on Form
            DC-2 and Sections II and III, and stamp each page with a
            consecutive number; however, do not number Form DC-2.  Inventory
            the JSF by reviewing the document numbers and recording page
            number ranges in the columns provided on Form DC-2.  The
            Contractor shall verify and record in the "Comments" section on
            Form DC-2 all intentional gaps in the page numbering sequence (for
            example,  "page numbers not used, XXXX - XXXX, XXXX - XXXX.  If
            there are no documents for a specific document type, enter a "NA"
            in the empty space.

3.20.2.2    Certain laboratory-specific documents related to the CSF may not
            fit into a clearly defined category.  The Contractor shall review
            Form DC-2 to determine if it is most appropriate to place them
            under categories 7, 8, 9,  or 10.  Category 10 should be used if
            there is no appropriate previous category.  These types of
            documents should be described or listed in the blanks under each
            appropriate category on Form DC-2.

3.20.2.3    If it is necessary to insert new or inadvertently omitted
            documents,  the Contractor shall identify the documents with unique
            accountable numbers and record the unique accountable numbers and
            the locations of the documents in the CSF in the "Other Records"
            section on Form DC-2.
                                     B-62                              OLM03.0

-------
                                                        Exhibit B — Section 4
                                                          Data Reporting Forms
4.0   DATA REPORTING FORMS

      The data reporting forms are shown on the following pages.
                                     B-63                              OLM03.0

-------
                                       1A
                     VOLATILE ORGANICS ANALYSIS DATA SHEET
                                 EPA SAMPLE NO.
Lab Name:

Lab Code:
                    Case No.:
      Contract:_

       SAS No.:
                                                         SDG No.:
Matrix: (soil/water)	

Sample wt/vol:      	(g/mL)

Level:   (low/med)  	

% Moisture: not dec.	

GC Column:  	 ID: 	

Soil Extract Volume: 	(uL)
               Lab Sample ID:

               Lab File ID:

               Date Received:

               Date Analyzed:
(mm)
Dilution Factor:
       CAS NO.
                       COMPOUND
               Soil Aliquot Volume:

         CONCENTRATION UNITS:
         (ug/L or ug/Kg)	
                          (uL)






•7 c nn 7 —



£7 — AA 1 —

1 C — 1 c; n





j- . « C.Q n
D **U J y U
A7_££— "5— — — .



-J Q Q T T

7 1 _c; c; _£; _ _ — .

c £_o "5 — c: _





i nn A i n i c; _,









i nnc.i no A

7 c_O c; _o — — _





1 O 7 T Q A— __..





i nn— an 7





i 77n— on 7 — — -




_ ,

_ _ . , -,, -, . ,

— ,-, -i . — <-*^4- Vi a nc>
c. n x o IT o s t n 3. n s


7\ y-»*-*4- x-^ ^.

—.i n • — ~\ f ' fin

-._.., ^ , ,

i -. —. • i -i , »



-,, -i — .^

-—_.., ^ . ,

« _ ,

- - - rp^.'—UI ^-v vs-vn4- U -, «^v

_, v,^ _, , ..._-..- u, 1 -. '^3x-v

_ J * U 1 *- U

T ") f* * *-,Vi T >-w^v-^v ^s.*^. «s

1 O T^ ' W 1

m • u T 4- u

._ . , , > , ,

--__,. ,1 ,,

_

, - ~j _. . , ,

D -C

yi V* j_ i_ I ^T* .*-

n n

_, . i > » ,

--«O(_. .-. ,,

m i

-,, -i ,

__ , , ^ ,

_ ,










































































































































                                  FORM I VOA
                                        OLM03.0

-------
                                      IB
                   SEMIVOLATILE ORGANICS  ANALYSIS  DATA SHEET
                           EPA SAMPLE NO.
Lab Name:

Lab Code:
                    Case No.:
Contract:

 SAS No.:
                                                        SDG No.:
Matrix: (soil/water)	

Sample wt/vol:      	(g/mL)	

Level:   (low/med)  	

% Moisture:  	  decanted: (Y/N)	

Concentrated Extract Volume:  	(uL)

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	       pH:	
         Lab Sample ID:

         Lab File ID:

         Date Received:

         Date Extracted:

         Date Analyzed:
         Dilution Factor:
       CAS NO.
                       COMPOUND
   CONCENTRATION UNITS:
   (ug/L or ug/Kg)	
       108-95-2	Phenol	
       111-44-4	bis ( 2-Chloroethyl) ether	
       95-57-8	2-Chlorophenol	
       541-73-1	1, 3-Dichlorobenzene	
       106-46-7	1, 4-Dichlorobenzene	
       95-50-1	1, 2-Dichlorobenzene	
       95-48-7	2-Methylphenol	
       108-60-1	2,2 '-oxybis( 1-Chloropropane)
       106-44-5	4-Methylphenol	
       621-64-7	N-Nitroso-di-n-propylamine	
       67-72-1	Hexachloroethane	
       98-95-3	Nitrobenzene	
       78-59-1	Isophorone	
       88-75-5	2-Nitrophenol	
       105-67-9	2, 4-Dimethylphenol	
       111-91-1	bis(2-Chloroethoxy)methane	
       120-83-2	2 , 4-Dichlorophenol	
       120-82-1	1, 2 , 4-Trichlorobenzene	
       91-20-3	Naphthalene	
       106-47-8	4-Chloroaniline	
       87-68-3	Hexachlorobutadiene	
       59-50-7	4-Chloro-3-methylphenol	
       91-57-6	2-Methylnaphthalene	
       77-47-4	Hexachlorocyclopentadiene	
       88-06-2	2,4,6-Trichlorophenol	
       95-95-4	2 , 4, 5-Trichlorophenol	
       91-58-7	2-Chloronaphthalene	
       88-74-4	2-Nitroaniline	
       131-11-3	Dimethylphthalate	
       208-96-8	Acenaphthylene	
       606-20-2	2 , 6-Dinitrotoluene	
       99-09-2	3-Nitroaniline	
       83-32-9	Acenaphthene	
                                  FORM  I  SV-1
                                 OLM03.0

-------
                                      1C
                   SEMIVOLATILE ORGANICS ANALYSIS DATA SHEET
                                            EPA SAMPLE NO.
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Matrix: (soil/water)	

Sample wt/vol:      	(g/mL)

Level:   (low/med)  	

% Moisture:  	  decanted: (Y/N)

Concentrated Extract Volume:  	

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	       pH:
                          Lab Sample ID:

                          Lab File ID:

                          Date Received:

                          Date Extracted:

                          Date Analyzed:
       CAS NO.
                       COMPOUND
                          Dilution Factor:
                    CONCENTRATION UNITS:
                    (ug/L or ug/Kg)	
       51-28-5	2 , 4-Dinitrophenol	
       100-02-7	4-Nitrophenol	
       132-64-9	Dibenzofuran	
       121-14-2	2 , 4-Dinitrotoluene	
       84-66-2	Diethylphthalate	
       7005-72-3	4-Chlorophenyl-phenylether
       86-73-7	Fluorene	
       100-01-6	4-Nitroaniline	
       534-52-1	4, 6-Dinitro-2-methylphenol_
       86-30-6	N-Nitrosodiphenylamine (1)
       101-55-3	4-Bromophenyl-phenylether	
       118-74-1	Hexachlorobenzene	
       87-86-5	Pentachlorophenol	
       85-01-8	Phenanthrene	
       120-12-7	Anthracene	
       86-74-8	Carbazole	
       84-74-2	Di-n-butylphthalate	
       206-44-0	Fluoranthene	
       129-00-0	Pyrene	
       85-68-7	Butylbenzylphthalate	
       91-94-1	3, 3 ' -Dichlorobenzidine	
       56-55-3	Benzo(a) anthracene	
       218-01-9	Chrysene	
       117-81-7	bis ( 2-Ethylhexyl) phthalate_
       117-84-0	Di-n-octylphthalate	
       205-99-2	Benzo(b) f luoranthene	
       207-08-9	Benzo(k) f luoranthene	
       50-32-8	Benzo (a) pyrene	
       193-39-5	Indeno(l, 2, 3-cd)pyrene	
       53-70-3	Dibenz (a, h) anthracene	
       191-24-2	Benzo(g,h, i)perylene	
     (1)  - Cannot be separated from Diphenylamine
                                  FORM  I  SV-2
                                                  OLM03.0

-------
                                      ID
                    PESTICIDE ORGANICS ANALYSIS DATA SHEET
                                  EPA SAMPLE  NO.
Lab Name:

Lab Code:
                    Case No.:
       Contract:

        SAS No.:
                                                        SDG No.:
Matrix:  (soil/water)	

Sample wt/vol:      	(g/mL)	

% Moisture:  	  decanted:  (Y/N)
Extraction:   (SepF/Cont/Sonc)

Concentrated Extract Volume:

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	
      (uL)
Lab Sample ID:

Lab File ID:

Date Received:

Date Extracted:

Date Analyzed:
                Dilution Factor:
PH:_
       CAS NO.
                       COMPOUND
Sulfur Cleanup: (Y/N)
          CONCENTRATION UNITS:
          (ug/L or ug/Kg)	




"31 Q Pft_Q_ ____



"7ft AA— P— — — —


ouy uu— £. —




ftn c^ *7— i — — —

"7O c; c Q _ — — _

*7O_ OH — P — —

7 7O 1 7 Ac; Q — _



1 H 7 1 H7 Q — —



"7*5 A "5 £ — —



*7 AO 1 — Q 7 A





onn i T t o



111 O A — *5 P O



COAftQ O1 Q— —

10^*7*5 — 00 — ft— —

1 1 r\Q 1 ftQ 1 — —

1 1 HQ ft PO c; — —













Alarm




— . . i j



— » .











_ , , .



1 W *-«W 1 y-1

c^l 1 <3

m U

_ i inn/"



_ -• 1 O "? O




























































































































                                 FORM I PEST
                                                                       OLM03.0

-------
                                       IE
                     VOLATILE ORGANICS ANALYSIS DATA SHEET
                       TENTATIVELY  IDENTIFIED  COMPOUNDS
                                 EPA SAMPLE NO.
Lab Name:

Lab Code:
                    Case No.:
      Contract:

       SAS No.:
                                                         SDG No.:
Matrix: (soil/water)	

Sample wt/vol:      	(g/mL)

Level:   (low/med)  	

% Moisture: not dec.	

GC Column: 	 ID: 	

Soil Extract Volume: 	(uL)
               Lab Sample ID:

               Lab File ID:

               Date Received:

               Date Analyzed:
(mm)
Dilution Factor:
 Number TICs found:
               Soil Aliquot Volume:

         CONCENTRATION UNITS:
         (ug/L or ug/Kg)	
                          (uL)
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.

COMPOUND NAME































RT































EST. CONC.































Q































                                FORM I VOA-TIC
                                                                       OLM03.0

-------
                                      IF
                   SEMIVOLATILE ORGANICS ANALYSIS DATA SHEET
                       TENTATIVELY IDENTIFIED COMPOUNDS
                                                                EPA  SAMPLE  NO.
Lab Name:

Lab Code:
                    Case No.:
                                     Contract:

                                      SAS No.:
          SDG No.:
                           _(g/mL)_
Matrix: (soil/water)	

Sample wt/vol:      	

Level:   (low/med)  	

% Moisture:  	  decanted: (Y/N)_

Concentrated Extract Volume: 	

Injection Volume:   	(uL)

GPC Cleanup:   (Y/N)	       pH:
                                    (uL)
Lab Sample ID:

Lab File ID:

Date Received:

Date Extracted:_

Date Analyzed:
                                              Dilution Factor:
 Number TICs found:
                                        CONCENTRATION UNITS:
                                        (ug/L or ug/Kg)	
CAS NUMBER
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.

COMPOUND NAME































RT































EST. CONC.































Q































                                 FORM I SV-TIC
                                                                        OLM03.0

-------
                                      2A
              WATER VOLATILE SYSTEM MONITORING COMPOUND RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
                                    SDG No.

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























SMC1
( TOL ) #






























SMC2
(BFB)#






























SMC3
(DCE)#






























OTHER






























TOT
OUT






























                                                   QC LIMITS
               SMC1  (TOL) = Toluene-d8              (88-110)
               SMC2  (BFB) = Bromofluorobenzene      (86-115)
               SMC3  (DCE) = l,2-Dichloroethane-d4   (76-114)

               # Column to be used to flag recovery values

               * Values outside of contract required QC  limits
page 	 of 	
                                 FORM II VOA-1
                                                   OLM03.0

-------
                                      2B
               SOIL VOLATILE SYSTEM MONI1ORING COMPOUND RECOVERY
Lab Name:

Lab Code:
Level:(low/med)
                    Case No.:
Contract:

 SAS No.:
                                                        SDG No.
page
        of
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.





























SMC1
(TOL)#





























SMC2
(BFB)#
3====:==





























SMC3
(DCE)#





























OTHER





























TOT
OUT





























                                                   QC LIMITS
               SMC1 (TOL) = Toluene-d8             (84-138)
               SMC2 (BFB) = Bromofluorobenzene     (59-113)
               SMC3 (DCE) = l,2-Dichloroethane-d4  (70-121)

               # Column to be used to flag recovery values

               * Values outside of contract required QC limits
                                 FORM II  VOA-2
                                  OLM03.0

-------
                                       2C
                     WATER  SEMIVOLATILE  SURROGATE RECOVERY
 Lab Name:

 Lab Code:
Case No.:
Contract:

 SAS No.:
                                                         SDG No.
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.





























SI
(NBZ)#





























S2
{ FBP ) #





























S3
(TPH)#





























S4
(PHL)#





























S5
(2FP)#





























S6
(TBP)#





























S7
(2CP)#





























SB
(DCB)#





























TOT
OUT





























                SI  (NBZ) = Nitrobenzene-d5
                S2  (FBP) = 2-Fluorobiphenyl
                S3  (TPH) = Terphenyl-dl4
                S4  (PHL) = Phenol-d5
                S5  (2FP) = 2-Fluorophenol
                S6  (TBP) = 2,4,6-Tribromophenol
                S7  (2CP) = 2-Chlorophenol-d4
                S8  (DCB) = 1,2-Dichlorobenzene-c
                # Column to be used to flag recovery values
                * Values outside of contract required QC  limits
                D Surrogate diluted out
QC LIMITS
(35-114)
(43-116)
(33-141)
(10-110)
(21-110)
(10-123)
(33-110)
(16-110)
(advisory)
(advisory)
page
        of
                                 FORM II SV-1
                                                                       OLM03.0

-------
                                      2D
                     SOIL SEMIVOLATILE SURROGATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
Level:(low/med)
                   SDG No.:
EPA
SAMPLE NO.






























SI
(NBZ)#






























S2
(FBP)#






























S3
( TPH ) #






























S4
(PHL)#






























S5
(2FP)#






























S6
(TBP)#






























S7
(2CP)#






























S3
(DCB)#






























TOT
OUT






























  01
  02
  03
  04
  05
  06
  07
  08
  09
  10
  11
  12
  13
  14
  15
  16
  17
  18
  19
  20
  21
  22
  23
  24
  25
  26
  27
  28
  29
  30
                SI (NBZ)
                S2 (FBP)
                S3 (TPH)
                S4 (PHL)
                S5 (2FP)
                S6 (TBP)
                S7 (2CP)
                S8 (DCB)
                              QC LIMITS
       Nitrobenzene-d5        (23-120)
       2-Fluorobiphenyl       (30-115)
       Terphenyl-dl4          (18-137)
       Phenol-d5              (24-113)
       2-Fluorophenol         (25-121)
       2,4,6-Tribromophenol   (19-122)
       2-Chlorophenol-d4      (20-130)
                       (advisory)
     = 1,2-Dichlorobenzene-d4 (20-130)  (advisory)
                # Column to be used to flag recovery values
                * Values outside of contract required QC limits
                D Surrogate diluted out
page 	 of 	
                                 FORM II SV-2
                                                  OLM03.0

-------
                                       2E
                       WATER PESTICIDE SURROGATE  RECOVERY
 Lab Name:

 Lab Code:
GC  Column(1):
Case No.:

     ID:
   Contract:

    SAS No.:
                                                         SDG No.:
(mm)   GC Column(2):
                                             ID:
(mm)

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























TCX 1
%REC #






























TCX 2
%REC #






























DCS 1
%REC #






























DCB 2
%REC #






























OTHER
(1)






























OTHER
(2)






























TOT
OUT






























           TCX = Tetrachloro-m-xylene
           DCB = Decachlorobiphenyl
                    QC  LIMITS

                    (30-150)
                    (30-150)
           # Column to be used to flag recovery values
           * Values outside of QC limits
           D Surrogate diluted out
page
        of
                                FORM II PEST-1
                                                   OLM03.0

-------
                                      2F
                       SOIL PESTICIDE  SURROGATE  RECOVERY
Lab Name:

Lab Code:
GC Column(l):
Case No.:

     ID:
                 Contract:_

                  SAS No.:
                                    SDG No.:
(mm)   GC Column(2):
                                             ID:
(mm)
EPA
SAMPLE NO.






























TCX 1
%REC #






























TCX 2
%REC #






























DCB 1
%REC #






























DCB 2
%REC #






























OTHER
(1)






























OTHER
(2)






























TOT
OUT






























       01
       02
       03
       04
       05
       06
       07
       08
       09
       10
       11
       12
       13
       14
       15
       16
       17
       18
       19
       20
       21
       22
       23
       24
       25
       26
       27
       28
       29
       30
           TCX = Tetrachloro-m-xylene
           DCB = Decachlorobiphenyl
                    QC LIMITS

                    (30-150)
                    (30-150)
           # Column to be used to flag recovery values
           * Values outside of QC limits
           D Surrogate diluted out
page
        of
                                FORM II PEST-2
                                                    OLM03.0

-------
                                       3A
          WATER VOLATILE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
 Lab  Name:

 Lab  Code:
                     Case No.:
Contract:

 SAS No.:
                                                         SDG No.:
Matrix  Spike  - EPA Sample No.:
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene

SPIKE
ADDED
(ug/L)






SAMPLE
CONCENTRATION
(ug/L)






MS
CONCENTRATION
(ug/L)






MS
%
REC #






QC.
LIMITS
REC.
61-145
71-120
76-127
76-125
75-130

COMPOUND
1 , 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene

SPIKE
ADDED
(ug/L)





MSD
CONCENTRATION
(ug/L)





MSD
%
REC #





%
RPD #





QC L:
RPD
14
14
11
13
13
:MITS
REC.
61-145
71-120
76-127
76-125
75-130
# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits

RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits
COMMENTS:
                                FORM III VOA-1
                                                                       OLM03.0

-------
                                      3B
          SOIL VOLATILE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Matrix Spike - EPA Sample No.:
Contract:

 SAS No.:
SDG No.:
                          Level:(low/med)
COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene

SPIKE
ADDED
(ug/Kg)






SAMPLE
CONCENTRATION
(ug/Kg)






MS
CONCENTRATION
(ug/Kg)






MS
%
REC #






QC.
LIMITS
REC.
59-172
62-137
66-142
59-139
60-133

COMPOUND
1, 1-Dichloroethene
Trichloroethene
Benzene
Toluene
Chlorobenzene

SPIKE
ADDED
(ug/Kg)






MSD
CONCENTRATION
(ug/Kg)






MSD
%
REC #






%
RPD #






QC L:
RPD
22
24
21
21
21
[MITS
REC.
59-172
62-137
66-142
59-139
60-133
# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits

RPD:       out of 	 outside limits
Spike Recovery:	 out of 	 outside limits
COMMENTS:
                                FORM III VOA-2
                                                   OLM03.0

-------
                                      3C
        WATER SEMIVOLATILE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
                 Contract:

                  SAS No.:
SDG No.:
Matrix Spike - EPA Sample No.:
COMPOUND
Phenol
2-Chlorophenol
1 , 4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene

SPIKE
ADDED
(ug/L)












SAMPLE
CONCENTRATION
(ug/L)












MS
CONCENTRATION
(ug/L)












MS
%
REC #












QC.
LIMITS
REC.
12-110
27-123
36- 97
41-116
39- 98
23- 97
46-118
10- 80
24- 96
9-103
26-127

COMPOUND
Phenol
2-Chlorophenol
1 , 4-Dichloroben2ene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene

SPIKE
ADDED
(ug/L)











MSD
CONCENTRATION
(ug/L)











MSD
%
REC #











%
RPD #











QC L:
RPD
42
40
28
38
28
42
31
50
38
50
31
:MITS
REC.
12-110
27-123
36- 97
41-116
39- 98
23- 97
46-118
10- 80
24- 96
9-103
26-127
 (1) N-Nitroso-di-n-propylamine

# Column to be used to flag recovery and RPD values with an asterisk
* Values outside of QC limits

RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits

COMMENTS:
                                 FORM III  SV-1
                                                   OLM03.0

-------
                                      3D
        SOIL SEMIVOLATILE MATRIX SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
                 Contract:

                  SAS No.:
SDG No.:
Matrix Spike - EPA Sample No.:
                          Level:(low/med)
COMPOUND
Phenol
2-Chlorophenol
1 , 4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Tr ichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol
2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene

SPIKE
ADDED
(ug/Kg)












SAMPLE
CONCENTRATION
(ug/Kg)












MS
CONCENTRATION
{ ug/Kg)












MS
%
REC #












QC.
LIMITS
REC.
26- 90
25-102
28-104
41-126
38-107
26-103
31-137
11-114
28- 89
17-109
35-142

COMPOUND
Phenol
2-Chlorophenol
1 , 4-Dichlorobenzene
N-Nitroso-di-n-prop. (1)
1,2, 4-Trichlorobenzene
4-Chloro-3-methylphenol
Acenaphthene
4-Nitrophenol

2 , 4-Dinitrotoluene
Pentachlorophenol
Pyrene

SPIKE
ADDED
(ug/Kg)













MSD
CONCENTRATION
(ug/Kg)













MSD
%
REC #













%
RPD #













QC L:
RPD
35
50
27
38
23
33
19
50
AT

47
36

MITS
REC.
26- 90
25-102
28-104
41-126
38-107
26-103
31-137
11-114
O Q QQ

17-109
35-142

 (1) N-Nitroso-di-n-propylamine

# Column to be used to flag recovery and RPD values with an  asterisk
* Values outside of QC limits

RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits
COMMENTS:
                                 FORM III SV-2
                                                                        OLM03.0

-------
                                       3E
         WATER PESTICIDE MATRIX  SPIKE/MATRIX SPIKE DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
SDG No.:
Matrix Spike - EPA Sample No.:
COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT

SPIKE
ADDED
(ug/L)







SAMPLE
CONCENTRATION
(ug/L)







MS
CONCENTRATION
(ug/L)







MS
%
REC /







QC.
LIMITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127

COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT

SPIKE
ADDED
(ug/L)







MSD
CONCENTRATION
(ug/L)







MSD
%
REC #







%
RPD #







QC L:
RPD
15
20
22
18
21
27
:MITS
REC.
56-123
40-131
40-120
52-126
56-121
38-127
# Column to be used to flag recovery and RPD values with an asterisk

* Values outside of QC limits
RPD:	 out of 	 outside limits
Spike Recovery:	 out of 	 outside limits
COMMENTS:
                                FORM III  PEST-1
                                                   OLM03.0

-------
                                      3F
          SOIL PESTICIDE MATRIX SPIKE/MATRIX SPIKE  DUPLICATE RECOVERY
Lab Name:

Lab Code:
Case No.:
                 Contract:_

                  SAS No.:
SDG No.:
Matrix Spike - EPA Sample No.:
COMPOUND
qamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT

SPIKE
ADDED
(ug/Kg)







SAMPLE
CONCENTRATION
(ug/Kg)







MS
CONCENTRATION
(ug/Kg)







MS
%
REC #







QC.
LIMITS
REC.
46-127
35-130
34-132
31-134
42-139
23-134

COMPOUND
qamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT

SPIKE
ADDED
(ug/Kg)







MSD
CONCENTRATION
(ug/Kg)







MSD
%
REC #







%
RPD #







QC L]
RPD
50
31
43
38
45
50
:MITS
REC.
46-127
35-130
34-132
31-134
42-139
23-134
# Column to be used to  flag recovery  and  RPD values with  an  asterisk

* Values outside of QC  limits
           out of 	 outside  limits
Spike Recovery:	 out of 	 outside  limits
COMMENTS:
                                FORM III PEST-2
                                                                        OLM03.0

-------
 Lab Name:

 Lab Code:
                                       4A
                         VOLATILE METHOD  BLANK  SUMMARY
                Contract:_

                 SAS No.:
                                           EPA SAMPLE NO.
                     Case No.:
                         SDG No.:
 Lab  File  ID:

 Date Analyzed:

 GC Column:
                         Lab Sample ID:

                         Time Analyzed:
ID:
(mm)
Heated Purge: (Y/N)
 Instrument  ID:
       THIS  METHOD  BLANK APPLIES TO THE FOLLOWING SAMPLES, MS AND MSD:
COMMENTS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
EPA
SAMPLE NO.






























LAB
SAMPLE ID






























LAB
FILE 10






























TIME
ANALYZED






























page 	 of 	
                                  FORM IV VOA
                                                                       OLM03.0

-------
Lab Name:

Lab Code:
               4B
SEMIVOLATILE METHOD BLANK SUMMARY


          	   Contract:	

               SAS No.:
                                                                EPA SAMPLE  NO.
Case No.:
SDG No.:
Lab File ID:      	

Instrument ID:

Matrix: (soil/water)

Level:(low/med)
                      Lab  Sample  ID:

                      Date Extracted:

                      Date Analyzed:

                      Time Analyzed:
      THIS METHOD BLANK APPLIES TO THE FOLLOWING SAMPLES, MS AND MSD:
COMMENTS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30


EPA
SAMPLE NO.
































LAB
SAMPLE ID
































LAB
FILE ID
































DATE
ANALYZED

































































page 	 of 	
                                  FORM IV sv
                                                                        OLM03.0

-------
Lab Name:

Lab Code:
                                       4C
                         PESTICIDE  METHOD  BLANK  SUMMARY
   Contract:

    SAS No.:
                              EPA SAMPLE NO.
                    Case  No.:
                                                         SDG  No.:
Lab Sample ID:
      Lab File ID:
Matrix:(soil/water)
Sulfur Cleanup:  (Y/N)

Date Analyzed  (1): 	

Time Analyzed  (1): 	

Instrument ID  (1): 	

GC Column  (1): 	
      Extraction:(SepF/Cont/Sonc)

      Date Extracted:        	

      Date Analyzed (2):     	

      Time Analyzed (2):     	

      Instrument  ID (2):     	

(mm)   GC Column (2): 	
ID:
        (mm)
         THIS METHOD BLANK APPLIES TO THE  FOLLOWING  SAMPLES,  MS  AND  MSD:
COMMENTS:

01
02
03
04
05
06
07
08
09
10
11
12
13
.14
15
16
17
18
19
20
21
22
23
24
25
26
EPA
SAMPLE NO.


























LAB
SAMPLE ID


























DATE
ANALYZED 1


























DATE
ANALYZED 2


























page 	 of 	
                                 FORM IV PEST
                                     OLM03.0

-------
                                      5A
                 VOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
                           BROMOFLUOROBENZENE  (BFB)
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
                                    SDG No.:
Lab File ID:

Instrument ID:

GC Column:
 ID:
           (mm)
      BFB Injection Date:

      BFB Injection Time:

      Heated Purge: (Y/N)
m/e
50
75
95
96
173
174
175
176
177

ION ABUNDANCE CRITERIA
8.0 - 40.0% of mass 95
30.0 - 66.0% of mass 95
Base peak, 100% relative abundance
5.0 - 9.0% of mass 95
Less than 2.0% of mass 174
50.0 - 120.0% of mass 95
4.0 - 9.0 % of mass 174
93.0 - 101.0% of mass 174
5.0 - 9.0% of mass 176

% RELATIVE
ABUNDANCE




( )1

( )1
( )1
( )2

        1-Value is % mass 174
                          2-Value is % mass 176
THIS CHECK APPLIES TO THE FOLLOWING SAMPLES, MS, MSD, BLANKS, AND  STANDARDS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
EPA
SAMPLE NO.






















LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















page 	 of 	
                                  FORM V VOA
                                                                        OLM03.0

-------
                                       5B
               SEMIVOLATILE ORGANIC INSTRUMENT PERFORMANCE CHECK
                     DECAFLUOROTRIPHENYLPHOSPHINE  (DFTPP)
Lab Name:

Lab Code:
Case No.:
                 Contract:_

                  SAS No.:
SDG No.:
Lab File ID:
Instrument ID:
                     DFTPP Injection Date:

                     DFTPP Injection Time:
m/e
51
68
69
70
127
197
198
199
275
365
441
442
443

ION ABUNDANCE CRITERIA
30.0 - 80.0% of mass 198
Less than 2.0% of mass 69
Mass 69 relative abundance
Less than 2.0% of mass 69
25.0 - 75.0% of mass 198
Less than 1.0% of mass 198
Base Peak, 100% relative abundance
5.0 to 9.0% of mass 198
10.0 - 30.0% of mass 198
Greater than 0.75% of mass 198
Present, but less than mass 443
40.0 - 110.0% of mass 198
15.0 - 24.0% of mass 442

% RELATIVE
ABUNDANCE

( )1

( )1








( )2

        1-Value is % mass 69
                          2-Value is % mass 442
THIS CHECK APPLIES TO THE FOLLOWING SAMPLES, MS, KSD,  BLANKS,  AND  STANDARDS:

01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
EPA
SAMPLE NO.






















LAB
SAMPLE ID






















LAB
FILE ID






















DATE
ANALYZED






















TIME
ANALYZED






















page
        of
                                   FORM V SV
                                                   OLM03.0

-------
                                      6A
                  VOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Case No.:
Contract:_

 SAS No.:
SDG No.
Instrument ID:
Heated Purge: (Y/N) 	

GC Column:           ID:
       Calibration Cate(s):

       Calibration Times:

      	(mm)
LAB FILE ID: RRF10 =
RRF20 =
RRF50 - RRF100- RRF200=

COMPOUND
Chloromethane
RRF10

Bromomethane *
Vinyl Chloride *
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide




1, 1-Dichloroethene *
1, 1-Dichloroethane *
1,2-Dichloroethene (total)

Chloroform *
1,2-Dichloroethane *
2-Butanone

1,1, 1-Trichloroethane *
Carbon Tetrachloride *
Bromodichloromethane *
1, 2-Dichloropropane

cis-1, 3-Dichloropropene *
Trichloroethene *
Dibromochloromethane *
1, 1,2-Trichloroethane *
Benzene *
trans-1, 3-Dichloropropene *
Bromoform *
4-Methyl-2-Pentanone
2-Hexanone


Tetrachloroethene *
1, 1,2,2-Tetrachloroethane *
Toluene *
Chlorobenzene *
Ethylbenzene *
Styrene *
Xylene (total) *
Toluene-d8

Bromof luorobenzene *
1, 2-Dichloroethane-d<



RRF20





































RRF50





































RRF100





































RRF200





































RRF






































%
RSD

*
*




*
*

*
*

*
*
*

*
*
*
*
*
*
*


*
*
*
*
*
*
*

*


* Compounds with required minimum RRF and maximum %RSD values.
  All other compounds must meet a minimum RRF of 0.010.
                                  FORM VI VOA
                                                                       OLM03.0

-------
                                      6B
                SEMIVOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Instrument ID:
                    Case No.:
          Contract:

           SAS No.:
Calibration Date(s):

Calibration Times:
                                                        SDG No.:
LAB FILE ID: RRF20 = RRF50 =
RRF80 = RRF120= RRF160=

COMPOUND RRF20
Phenol *
bis (2-Chloroethyl) ether *
2-Chlorophenol *
1, 3-Dichlorobenzene *
1, 4-Dichlorobenzene *
1, 2-Dichlorobenzene *
2-Methylphenol *
2,2' -oxybis ( 1-Chloropropane )
4-Methylphenol *
N-Nitroso-di-n-propylamine *
Hexachloroethane *
Nitrobenzene *
Isophorone *
2-Nitrophenol *
2, 4-Dimethylphenol *
bis(2-Chloroethoxy)methane *
2, 4-Dichlorophenol *
1, 2 , 4-Trichlorobenzene *
Naphthalene *
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-methylphenol *
2-Methylnaphthalene *
Hexachlorocyclopentadiene
2 , 4, 6-Trichlorophenol *
2,4, 5-Trichlorophenol *
2-Chloronaphthalene *
2-Nitroaniline
Dimethylphthalate
Acenaphthylene *
2 , 6-Dinitrotoluene *
3-Nitroaniline
Acenaphthene *
2 , 4-Dinitrophenol
4-Nitrophenol
Dibenzofuran *
2,4-Dinitrotoluene *

RRF50






































RRF80






































RRF120






































RRF160







































RRF







































%
RSD
*
*
*
*
*
*
•*

*
*
*
it
*
*
*
*
*
*
*


*
*
|
*
*
*


*
*

*


*
*

 Compounds with  required  minimum RRF  and  maximum %RSD  values.
 All other compounds must meet  a minimum  RRF of  0.010.
                                 FORM  VI  SV-1
                                           OLM03.0

-------
                                      6C
                SEMIVOLATILE ORGANICS INITIAL CALIBRATION DATA
Lab Name:

Lab Code:
Instrument ID:
Case No.:
Contract:

 SAS No.:
                                                        SDG No.
       Calibration Date(s):

       Calibration Times:
LAB FILE ID: RRF20 =
RRF80 = RRF120=


COMPOUND
Diethylphthalate
4-Chlorophenyl-phenylether '
Fluorene <
4-Nitroaniline
4 , 6-Dinitro-2-methy Iphenol
N-Nitrosodiphenylamine (1)
4-Bromophenyl-phenylether '
Hexachlorobenzene *
Pentachlorophenol i
Phenanthrene <
Anthracene i
Carbazole
Di-n-butylphthalate
Fluoranthene "
Pyrene y
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo( a) anthracene y
Chrysene *
bis (2-Ethylhexyl ) phthalate
Di-n-octylphthalate
Benzo(b) f luoranthene '
Benzo(k) f luoranthene *
Benzo(a)pyrene i
Indeno(l,2,3-cd)pyrene <
Dibenz (a, h) anthracene i
Benzo(g,h, i)perylene '

Nitrobenzene-d5
2-Fluorobiphenyl i
Terphenyl-dl4 i
Phenol-d5 '
2-Fluorophenol i
2,4, 6-Tribromophenol
2-Chlorophenol-d4 '
l,2-Dichlorobenzene-d4 '





RRF20

it
*



t
t
t
Ir
t


c
e


t
t


t
t
\e
t
t
t


t
*
te
it

t
It





RRF50





































RRF50
RRF16C


RRF80





































=
)=


RRF120









































RRF160









































RRF








































%
RSD

*
*



*
A
A
*
A


A
A


A
A


A
A
A
A
A
A

A
A
A
4
A

*
A

(1) Cannot be separated from Diphenylamine
* Compounds with required minimum RRF and maximum %RSD values.
  All other compounds must meet a minimum RRF of 0.010.
                                 FORM VI SV-2
                                                   OLM03.0

-------
                                      6D
          PESTICIDE INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES
Lab Name:

Lab Code:
                    Case No.:
Contract:

 SAS No.:
                                                        SDG No.:
Instrument ID:

GC Column:
               mid
	  Level  (x  low):  low 	

 ID: 	(mm)   Date(s) Analyzed:
high
COMPOUND
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4'-DDE
Endrin
Endosulfan II
4, 4 '-ODD
Endosulfan sulfate
4, 4 '-DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane

Tetrachloro-m-xylene
Decachlorobiphenyl

RT Ol
LOW
























? STAND;
MID
























1RDS
HIGH
























MEAN
RT
























RT w:
FROM
























:NDOW
TO
























     *  Surrogate retention times are measured from Standard Mix A analyses.

     Retention time windows are + 0.05 minutes for all compounds that elute
     before Heptachlor epoxide,  +0.07 minutes for all other compounds,
     except iO.10 minutes for Decachlorobiphenyl.
                                FORM VI  PEST-1
                                                                       OLM03.0

-------
                                      6E
          PESTICIDE INITIAL CALIBRATION OF SINGLE COMPONENT ANALYTES
Lab Name:

Lab Code:
Instrument ID:

GC Column:
Case No.:
                Contract:

                 SAS No.:
SDG No.:
                                mid
	  Level  (x  low):  low

 ID: 	(mm)   Date(s) Analyzed:
      high
COMPOUND
alDha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4, 4 '-DDE
Endrin
Endosulfan II
4, 4 '-ODD
Endosulfan sulfate
4, 4 '-DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane

Tetrachloro-m-xylene
Decachlorobiphenyl

LOW
























CALIBRATIC
MID
























>N FACTORS
HIGH
























MEAN
























%RSD
























* Surrogate calibration factors are measured from Standard Mix A analyses.
                                FORM VI PEST-2
                                                   OLM03.0

-------
                                      6F
           PESTICIDE INITIAL CALIBRATION OF MULTICOMPONENT ANALYTES
Lab Name:

Lab Code:
Case No.
                 Contract:

                  SAS No.:
SDG No.
Instrument ID: 	

GC Column:            ID:
                Date(s) Analyzed:
           (mm)
COMPOUND
Toxaphene
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
AMOUNT
(ng)
















PEAK
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
*1
*2
*3
4
5
RT









































RT W]
FROM









































NDOW
TO









































CALIBRATION
FACTOR









































        Denotes required peaks
                                 FORM VI  PEST-3
                                                                        OLM03.0

-------
                                      6G
                     PESTICIDE ANALYTE RESOLUTION SUMMARY
Lab Name:

Lab Code:
Case No.:
    Contract:

     SAS  No.:
SDG No.
GC Column  (I):
EPA Sample No. (Standard 1):

Date Analyzed  (1): 	
GC Column (2):
      ID:
EPA Sample No. (Standard 2):

Date Analyzed (2): 	
              _(mm)  Instrument ID (1):

              	 Lab Sample ID (1):

                    Time Analyzed (1):

01
02
03
04
05
06
07
08
09
ANALYTE









RT









RESOLUTION
(*)









 (mm)   Instrument  ID  (2):

	  Lab Sample  ID  (2):

       Time Analyzed  (2):

01
02
03
04
05
06
07
08
09
ANALYTE









RT









RESOLUTION
(%)









                                FORM VI PEST-4
                                                   OLM03.0

-------
                                       6H
                     PERFORMANCE EVALUATION MIXTURE  (PEM)
Lab Name:

Lab Code:
                    Case No.:
           Contract:

            SAS No.:
                      SDG No.
GC Column  (1):
ID:
EPA Sample No.  (Standard 1):

Date Analyzed  (1): 	
GC Column (2):
EPA Sample No. (Standard 2):

Date Analyzed (2): 	
 (mm)   Instrument  ID  (1):

	  Lab Sample  ID  (1):

       Time Analyzed  (1):

01
02
03
04
05
06
07
08
ANALYTE








RT








RESOLUTION
(%)








        (mm)   Instrument ID (2):

       	 Lab Sample ID (2):

              Time Analyzed (2):

01
02
03
04
05
06
07
08
ANALYTE








RT








RESOLUTION
(%)








                                FORM VI PEST-5
                                                                       OLM03.0

-------
Lab Name:

Lab Code:
                   61
     INDIVIDUAL STANDARD MIXTURE A


	   Contract:	

 Case No.:          SAS No.:
                                                         SDG No.
GC Column  (1):
       ID:
               (mm)   Instrument ID (1):
' N(
rzec


01
02
03
04
05
06
07
08
09
10
11
3. (Standard 1) : ]
3 (1):

ANALYTE











..ab Sam]
rime An;

RT











Die ID ( 1 ) :
ilyzed (1):

RESOLUTION
(%)


























GC Column (2):
               (mm)   Instrument  ID  (2):
No. (Standard 2): Lab Sample ID (2):
zed (2): Time Analyzed (2):
01
02
03
04
05
06
07
08
09
10
11

"ANALYTE












RT











RESOLUTION
(%)












                                FORM VI PEST-6
                                                    OLM03.0

-------
                                       6J
                         INDIVIDUAL STANDARD MIXTURE B
Lab Name:_

Lab Code:
Case No.:
                 Contract:

                  SAS No.:
                       SDG  No.
GC Column  (1):
                          ID:
EPA Sample No.  (Standard 1):

Date Analyzed  (1): 	
              (mm)  Instrument ID (1):

              	 Lab Sample ID (1):

                    Time Analyzed (1):

01
02
03
04
05
06
07
08
09
10
11
12
13
ANALYTE













RT













RESOLUTION
(%)













GC Column (2) :
      ID:
EPA Sample No. (Standard 2):

Date Analyzed (2): 	
 (mm)   Instrument  ID  (2):

	  Lab Sample  ID  (2):

       Time Analyzed  (2):

01
02
03
04
05
06
07
08
09
10
11
12
13
ANALYTE













RT













RESOLUTION
(*)













                                FORM VI PEST-7
                                                   OLM03.0

-------
                                      7A
                     VOLATILE CONTINUING CALIBRATION CHECK
Lab Name:

Lab Code:
                    Case No.:
          Contract:

           SAS  No.:
Instrument ID:

Lab File ID:
 Calibration  Date:
SDG No.

 Time:
Heated Purge: (Y/N) 	

GC Column:           ID:
 Init.  Calib.  Date(s):

 Init.  Calib.  Times:

	(mm)

COMPOUND
Chloromethane
Bromomethane
Vinyl Chloride
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide
1, 1-Dichloroethene
1 , 1-Dichloroethane
1, 2-Dichloroethene (total)
Chloroform
1 , 2-Dichloroethane
2-Butanone
1,1, 1-Trichloroethane
Carbon Tetrachloride
Bromodichloromethane
1, 2-Dichloropropane
cis-1, 3-Dichloropropene
Trichloroethene
Dibromochloromethane
1,1, 2-Trichloroethane
Benzene
trans-1 , 3-Dichloropropene
Bromoform
4-Methy 1-2 -Pent anone
2-Hexanone
Tetrachloroethene
1,1,2, 2-Tetrachloroethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Xylene (total)
Toluene-d8
Bromof luorobenzene
1, 2-Dichloroethane-d4


RRF

































======




RRF 50

































= =^ = =::= =



MIN
RRF

0.100
0.100




0.100
0.200

0.200
0.100

0.100
0.100
0.200

0.200
0.300
0.100
0.100
0.500
0.100
0.100


0.200
0.300
0.400
0.500
0.100
0.300
0.300
= = ~ = =
0.200



%D

































=~~^=^



MAX
%D

25.0
25.0




25.0
25.0

25.0
25.0

25.0
25.0
25.0

25.0
25.0
25.0
25.0
25.0
25.0
25.0


25.0
25.0
25.0
25.0
25.0
25.0
25.0
====
25.0


      All other compounds must meet a minimum RRF of 0.010.
                                 FORM VII VOA
                                                                       OLM03.0

-------
                                7B
             SEMIVOLATILE  CONTINUING  CALIBRATION CHECK
Lab Name:
Lab Code:
Instrument ID:
Lab File ID:

Case No. :
Contract:
SAS No . : SDG No . :
Calibration Date: Time:
Init.
Calib. Date(s) :
                      Init. Calib. Times:

COMPOUND
Phenol
bis ( 2-Chloroethyl ) ether
2 -Chlorophenol
1 , 3-Dichlorobenzene
1 , 4-Dichlorobenzene
1 , 2-Dichlorobenzene
2-Methylphenol
2,2' -oxybis ( 1-Chloropropane )
4-Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Tsophorone
2-Nitrophenol
2 , 4-Dimethylphenol
bis ( 2 -Chloroethoxy ) methane
2 , 4-Dichlorophenol
1,2, 4-Trichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-methylphenol
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2,4, 5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
Acenaphthylene
2 , 6-Dinitrotoluene
3-Nitroaniline
Acenaphthene
2 , 4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 , 4-Dinitrotoluene


RRF







































RRF50






































MIN
RRF
0.800
0.700
0.800
C.600
0.500
0.400
0.700

0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300
0.200
0.200
0.700


0.200
0.400

0.200
0.200
0.800


0.900
0.200

0.900


0.800
0.200


%D






































MAX
%D
25.0
25.0
25.0
25.0
25.0
25.0
25.0

25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0
25.0


25.0
25.0

25.0
25.0
25.0


25.0
25.0

25.0


25.0
25.0

All other compounds must meet a minimum RRF of 0.010.
                           FORM VII SV-1
OLM03.0

-------
                                  7C
              SEMIVOLATILE CONTINUING CALIBRATION CHECK
Lab Name:
Lab Code :
Instrument ID:
Lab File ID:

Case No. :
Contract:
SAS No. : SDG No. :
Calibration Date: Time:
Init.
Calib. Date(s):
                        Init.  Calib.  Times:

COMPOUND
Diethylphthalate
4-Chlorophenyl-phenylether
Fluorene
4-Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine (1)
4-Bromophenyl-phenylether
Hexachlorobenzene
Pentachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo ( a ) anthracene
Chrysene
bis (2-Ethylhexyl)phthalate
Di-n-octylphthalate
Benzo ( b ) f luoranthene
Benzo (k) f luoranthene
Benzo (a) pyrene
Indeno ( 1 , 2 , 3-cd ) pyrene
Dibenz ( a , h ) anthracene
Benzo (q,h, i)perylene

Nitrobenzene-d5
2-Fluorobiphenyl
Terphenyl-dl4
Phenol-d5
2-Fluorophenol
2,4, 6-Tribromophenol
2-Chlorophenol-d4
1 , 2-Dichlorobenzene-d4


RRF





































RRF50




































MIN
RRF
0.400
0.900



0.100
0.100
0.050
0.700
0.700


0.600
0.600


0.800
0.700


0.700
0.700
0.700
0.500
0.400
0.500

0.200
0.700
0.500
0.800
0.600

0.800
0.400


%D




































MAX
%D
25.0
25.0



25.0
25.0
25.0
25.0
25.0


25.0
25.0


25.0
25.0


25.0
25.0
25.0
25.0
25.0
25.0
	
25.0
25.0
25.0
25.0
25.0

25.0
25.0

(1)  Cannot be separated from Diphenylamine
  All other compounds must meet a minimum RRF of 0.010.
                             FORM  VII  SV-2
                                                                  OLM03.0

-------
                                      7D
                  PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name:
Lab Code:
GC Column:

Case No. :
ID:
Contract:
SAS No. : SDG No. :
(nun) Init. Calib. Date(s):
EPA Sample No.(PIBLK):

Lab Sample ID  (PIBLK):_

EPA Sample No.(PEM): _

Lab Sample ID  (PEM):
     Date Analyzed

     Time Analyzed

     Date Analyzed

     Time Analyzed
PEM
COMPOUND
alpha-BHC
beta-BHC
gamma-BHC (Lindane)
Endrin
4, 4 '-DDT
Methoxychlor

RT







RT W]
FROM







ENDOW
TO







CALC
AMOUNT
(ng)







NOM
AMOUNT
(ng)







%D







4,4'-DDT % breakdown (1):

Combined % breakdown (1):
Endrin % breakdown (1):
                               FORM VII PEST-1
                                                                       OLM03.0

-------
                                      7E
                  PESTICIDE CALIBRATION VERIFICATION SUMMARY
Lab Name:
Lab Code:
GC Column:

Case No. :
ID:
Contract:
SAS No. : SDG No. :
(mm) Init. Calib. Date(s):
EPA Sample No.(PIBLK):

Lab Sample ID (PIBLK):

EPA Sample No.(INDA):

Lab Sample ID (INDA):
Date Analyzed :

Time Analyzed :

Date Analyzed :_

Time Analyzed :
INDIVIDUAL MIX A
COMPOUND
alpha-BHC
qamma-BHC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4, 4 '-ODD
4, 4 '-DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl

RT












RT w:
FROM












:NDOW
TO












CALC
AMOUNT
(ng)












NOM
AMOUNT
(ng)












%D












EPA Sample No.(INDB)

Lab Sample ID (INDB)
Date Analyzed :

Time Analyzed :
INDIVIDUAL MIX B
COMPOUND
beta-BHC
delta-BHC
Aldrin
Heptachlor epoxide
4, 4 '-DDE
Endosulfan II
Endosulfan sulfate
Endrin ketone
Endrin aldehyde
alpha-Chlordane
aamma-Chlordane
Tetrachloro-m-xylene
Decachlorobiphenyl

RT













RT WI
FROM













NDOW
TO













CALC
AMOUNT
(ng)













NOM
AMOUNT
(ng)













%D













                                FORM VII PEST-2
                                                                       OLM03.0

-------
                                      8A
                VOLATILE INTERNAL STANDARD AREA AND RT SUMMARY
Lab Name:

Lab Code:
Case No.:
                 Contract:

                  SAS No.:
Lab File ID (Standard):

Instrument ID: 	

GC Column:           ID:
           (mm)
        SDG No.:
                            Date Analyzed:

                            Time Analyzed:
Heated Purge: (Y/N)

12 HOUR STD
UPPER LIMIT
LOWER LIMIT
EPA SAMPLE
NO.






















ISl(BCM)
AREA #


























RT #


























IS2(DFB)
AREA #


























RT #


























IS3(CBZ)
AREA #


























RT #


























 01
 02
 03
 04
 05
 06
 07
 08
 09
 10
 11
 12
 13
 14
 15
 16
 17
 18
 19
 20
 21
 22
    IS1 (BCM) = Bromochloromethane
    IS2 (DFB) = 1,4-Difluorobenzene
    IS3 (CBZ) = Chlorobenzene-d5

    AREA UPPER LIMIT = +100% of internal standard area
    AREA LOWER LIMIT = - 50% of internal standard area
    RT UPPER LIMIT = +0.50 minutes of internal standard RT
    RT LOWER LIMIT = -0.50 minutes of internal standard RT

    # Column used to flag values outside QC limits with an asterisk.
    * Values outside of QC limits.
page 	 of
                                 FORM VIII VOA
                                                                      OLM03.0

-------
                                      8B
              SEMIVOLATILE INTERNAL STANDARD AREA AND RT SUMMARY
Lab Name:

Lab Code:
Case No.:
Contract:

 SAS No.:
                                                        SDG No.
Lab File ID  (Standard):

Instrument ID:
                            Date Analyzed:

                            Time Analyzed:






01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22

12 HOUR STD
UPPER LIMIT
LOWER LIMIT

EPA SAMPLE
NO.






















ISl(DCB)
AREA #



























RT #



























IS2 (NPT)
AREA #



























RT #



























IS3(ANT)
AREA #



























RT #



























    IS1 (DCS) = l,4-Dichlorobenzene-d4
    IS2 (NPT) = Naphthalene-d8
    IS3 (ANT) = Acenaphthene-dlO

    AREA UPPER LIMIT = +100% of internal standard area
    AREA LOWER LIMIT = - 50% of internal standard area
    RT UPPER LIMIT = +0.50 minutes of internal standard RT
    RT LOWER LIMIT = -0.50 minutes of internal standard RT

    # Column used to flag internal standard area values with an asterisk.
    * Values outside of QC limits.

page    of
                                FORM VIII SV-1
                                                  OLM03.0

-------
                                      8C
              SEMIVOLATILE  INTERNAL STANDARD AREA AND RT SUMMARY
Lab Name:

Lab Code:
Case No.:
                 Contract:

                  SAS No.:
SDG No.
Lab File ID  (Standard):

Instrument ID:
                            Date Analyzed:

                            Time Analyzed:






01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22

12 HOUR STD
UPPER LIMIT
LOWER LIMIT

EPA SAMPLE
NO.






















IS4(PHN)
AREA #



























RT #



























ISS(CRY)
AREA #



























RT #



























IS6(PRY)
AREA #



























RT #



























    IS4 (PHN) = Phenanthrene-dlO
    IS5 (CRY) = Chrysene-dl2
    IS6 (PRY) = Perylene-dl2

    AREA UPPER LIMIT = +100% of internal standard area
    AREA LOWER LIMIT = - 50% of internal standard area
    RT UPPER LIMIT = +0.50 minutes of internal standard RT
    RT LOWER LIMIT = -0.50 minutes of internal standard RT

    # Column used to flag internal standard area values with an asterisk.
    * Values outside of QC limits.

page    of
                                FORM VIII SV-2
                                                  OLM03.0

-------
Lab Name:_

Lab Code:

GC Column:
                   8D
     PESTICIDE ANALYTICAL SEQUENCE

	   Contract:	

 Case No.:  	  SAS No.:  	  SDG No.:

   ID:  	(mm)  Init. Calib.  Date(s):	
Instrument ID:
  THE ANALYTICAL SEQUENCE OF PERFORMANCE EVALUATION MIXTURES, BLANKS,
               SAMPLES, AND STANDARDS IS GIVEN BELOW:
      01
      02
      03
      04
      05
      06
      07
      08
      09
      10
      11
      12
      13
      14
      15
      16
      17
      18
      19
      20
      21
      22
      23
      24
      25
      26
      27
      28
      29
      30
      31
      32
MEAN SURROGATE RT FROM INITIAL CALIBRATION
TCX: DCB:

EPA
SAMPLE NO.
































LAB
SAMPLE ID
































DATE
ANALYZED

































TIME
ANALYZED

































TCX
RT #
































DCB
RT #
































        TCX = Tetrachloro-m-xylene
        DCB = Decachlorobiphenyl
                     QC LIMITS
                  (+ 0.05 MINUTES)
                  (+ 0.10 MINUTES)
page
        # Column used to flag retention time values with  an  asterisk.
        * Values outside of QC  limits.

        of
                                FORM VIII PEST
                                                                       OLM03.0

-------
                                      9A
                      PESTICIDE FLORISIL CARTRIDGE CHECK
Lab Name:
                                     Contract:
Lab Code:
                    Case No. :
                                      SAS No.:
                                                        SDG No.:
Florisil Cartridge Lot Number:
        Date of Analysis:
GC Column(1):
                         ID:
(mm)   GC Column(2):
                                                                  ID:
(mm)
COMPOUND
alpha-BHC
gamma-BBC (Lindane)
Heptachlor
Endosulfan I
Dieldrin
Endrin
4, 4 '-ODD
4, 4 '-DDT
Methoxychlor
Tetrachloro-m-xylene
Decachlorobiphenyl

SPIKE
ADDED
(ng)












SPIKE
RECOVERED
(ng)












%
REC #












QC
LIMITS
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120
80-120

     # Column to be used to flag recovery with an asterisk.
     * Values outside of QC limits.

  THIS CARTRIDGE LOT APPLIES TO THE FOLLOWING SAMPLES, BLANKS, MS, AND MSD:
            01
            02
            03
            04
            05
            06
            07
            08
            09
            10
            11
            12
            13
            14
            15
            16
            17
            18
            19
            20
            21
            22
            23
EPA
SAMPLE NO.























LAB
SAMPLE ID























DATE
ANALYZED 1























DATE
ANALYZED 2























page
        of
                                FORM IX PEST-1
                                                                       OLM03.0

-------
                                      9B
                           PESTICIDE  GPC CALIBRATION
Lab Name:

Lab Code:
GPC Column:

GC Column(l):
                    Case No.:
            Contract:

             SAS No.:
                                                        SDG No.:
ID:
    Calibration Date:

(mm)   GC Column(2):
IDs
                                                  (mm)
COMPOUND
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT

SPIKE
ADDED
(ng)







SPIKE
RECOVERED
(ng)







%
REC #







QC.
LIMITS
REC.
80-110
80-110
80-110
80-110
80-110
80-110

          # Column to be used to flag recovery values with an asterisk
          * Values outside of QC limits

   THIS GPC CALIBRATION APPLIES TO THE FOLLOWING SAMPLES, BLANKS, MS AND MSD:
            01
            02
            03
            04
            05
            06
            07
            08
            09
            10
            11
            12
            13
            14
            15
            16
            17
            18
            19
            20
            21
            22
            23
            24
            25
            26
EPA
SAMPLE NO.


























LAB
SAMPLE ID


























DATE
ANALYZED 1


























DATE
ANALYZED 2


























page
        of
                                FORM IX PEST-2
                                             OLM03.0

-------
                                       10A
                        PESTICIDE IDENTIFICATION SUMMARY
                         FOR SINGLE COMPONENT ANALYTES
                                       EPA SAMPLE NO.
 Lab Name:

 Lab Code:
                     Case No.:
            Contract:

             SAS No.:
                                                         SDG No.:
 Lab  Sample ID :
 Instrument ID (1):

 GC  Column(l):  	
ID:
      Date(s)  Analyzed:

      Instrument ID (2):

(mm)   GC Column(2): 	
                                        ID:
                                                  (nun)
ANALYTE









COL
1
2
1
2
1
2
1
2
1
2
1
2
1
2
1
2
RT

















RT W
FROM

















:NDOW
TO

















CONCENTRATION

















%D









page
        of
                                 FORM X PEST-1
                                                                        OLM03.0

-------
                                       10B
                        PESTICIDE IDENTIFICATION SUMMARY
                           FOR MULTICOMPONENT ANALYTES
                                       EPA SAMPLE NO.
 Lab Name:

 Lab Code:
                     Case No.:
            Contract:

             SAS No.:
                                                         SDG No.:
 Lab Sample ID :
 Instrument  ID  (1):

 GC  Column(1):  	
ID:
      Date(s)  Analyzed:

      Instrument ID (2):

(mm)   GC Column(2): 	
                                        ID:
                                                  (mm)
ANALYTE

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2

COLUMN 1
COLUMN 2
PEAK
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
RT































RT W
FROM































:NDOW
TO































CONCENTRATION































MEAN
CONCENTRATION







%D




 At  least  3  peaks  for  each  column  are  required  for  identification  of
multicomponent analytes
page
        of
                                 FORM X PEST-2
                                                                        OLM03.0

-------
                                           SAMPLE LOG-IN SHEET
Lab Name
Received By (Print Name)
Page of

Log- in Date
Received By (Signature)
Case Number
Remarks:
l Custody Seal(s) Present/Absent*
Intact/Broken
2 Custody Seal Nos

3 Chain-of Custody Records Present/Absent*
4 Traffic Reports or Packing Lists Present/Absent*
5 Airbill Airbill/Sticker
Present/Absent*
6 Airbill No

7 Sample Tags Present/Absent*
Sample Tag Numbers Listed/Not Listed on
Chain-of-Custody
8 Sample Condition Intact/Sroken*/Leaking
9 Does information on custody Yes/No*
records, traffic reports, and
sample tags agree?
10 Date Received at Lab
11 Time Received

Sample Transfer
Fraction Fraction
Area # Area #
By By
on on
Sample Delivery Group No.
EPA Sample #

















Corresponding
Sample Tag #

















Assigned Lab it

















SAS Number
Remarks:
Condition of Sample
Shipment, etc

















*  Contact SMO and attach record of  resolution
Reviewed By
Date
Logbook NO
Logbook Page No
                                                FORM DC-1
                                                                                                     OLMO;

-------
                       ORGANICS  COMPLETE  SDG  FILE  (CSF)  INVENTORY SHEET
     LABORATORY NAME
     CITY/STATE 	
     CASE NO.  	 SDG NO.	 SDG NOS.  TO FOLLOW
     	              SAS  NO.
     CONTRACT NO.
     SOW NO.
     All  documents  delivered in  the  Complete SDG  File  must  be  original  documents
     where  possible.
                                                              PAGE NOs            CHECK
                                                             FROM   TO       LAB        EPA
1.  Inventory Sheet  (Form DC-2)  (Do not  number)
2.  SDG Case Narrative
3.  SDG Cover Sheet/Traffic Report
4.  Volatiles Data
    a.  QC Summary
       System Monitoring Compound Summary (Form II VOA)
       Matrix Spike/Matrix  Spike Duplicate Summary
            (Form III VOA)
       Method Blank Summary (Form IV  VOA)
       GC/MS  Instrument Performance Check (Form V VOA)
       Internal Standard Area and RT  Summary
            (Form VIII  VOA)

    b.  Sample Data
       TCL Results  - (Form  I  VOA)
       Tentatively  Identified Compounds  (Form I VOA-TIC)
       Reconstructed total  ion chromatograms (RIC)  for
         each sample
       For each sample:
         Raw spectra and background-subtracted mass
           spectra of  target compounds  identified
         Quantitation reports
         Mass spectra of all reported TICs with three
           best library matches

   c. Standards Data (All  Instruments)
       Initial  Calibration  Data (Form VI VOA)
      RICs and Quan Reports  for all  Standards
      Continuing Calibration Data  (Form VII VOA)
      RICs and Quantitation  Reports  for all Standards

   d. Raw QC Data
      BFB
      Blank Data
      Matrix Spike/Matrix  Spike Duplicate Data
                                         FORM DC-2-1                                    OLM03.0

-------
                    ORGANICS COMPLETE SDG FILE (CSF)  INVENTORY SHEET (Cent.)
     CASE NO.  	 SDG NO. 	 SDG NOS.  TO FOLLOW
                                SAS NO.
                                                              PAGE NOs            CHECK
                                                              FROM    TO        LAB       EPA
 5.  Semivolatiles Data
    a. QC  Summary
       Surrogate Percent  Recovery  Summary  (Form II SV)
       MS/MSD Summary  (Form  III  SV)
       Method Blank Summary  (Form  IV SV)
       GC/MS Instrument Performance Check  (Form V SV)
       Internal Standard  Area and  RT Summary
             (Form VIII SV)

    b. Sample Data
       TCL Results (Form  I SV-1, SV-2)
       Tentatively Identified Compounds  (Form I SV-TIC)
       Reconstructed total ion chromatograms (RIC) for
         each  sample
       For each sample:
         Raw spectra  and  background-subtracted mass
            spectra of target compounds
         Quantitation  reports
         Mass  spectra  of  TICs with three best library
            matches
         GPC chromatograms  (if GPC performed)

    c.  Standards Data (All Instruments)
       Initial  Calibration Data  (Form VI SV-1,  SV-2)
       RICs and Quan Reports for all Standards
       Continuing Calibration Data (Form VII SV-1, SV-2)
       RICs and Quantitation Reports for all Standards
   d.  Raw QC Data
       DFTPP
       Blank Data
       Matrix Spike/Matrix Spike Duplicate Data

   e.  Raw GPC Data

6.  Pesticides
   a.  QC Summary
       Surrogate Percent Recovery Summary (Form II  PEST)
       MS/MSD Duplicate Summary (Form III PEST)
       Method Blank Summary (Form IV PEST)
                                          FORM DC-2-2                                   OLM03.

-------
                   ORGANICS COMPLETE SDG FILE (CSF) INVENTORY SHEET (Cont.)
    CASE NO. 	 SDG'NO. 	 SDG NOS. TO FOLLOW
    	     SAS NO.
                                                              PAGE  NOs            CHECK
                                                             FROM   TO       LAB       EPA
6.  Pesticides  (Cont.)
   b.  Sample Data
       TCL Results -  Organic Analysis Data Sheet
            (Form I PEST)
       Chromatograms  (Primary Column)
       Chromatograms  from  second GC column confirmation
       GC Integration report or data system printout
       Manual work sheets
       For pesticides/Aroclors confirmed by GC/MS,
         copies of raw spectra and copies of background-
         subtracted mass spectra of target comoounds
         (samples & standards)

   c.  Standards  Data
       Initial  Calibration  of Single Component  Analytes
               (Form  VI PEST-1 and PEST-2)
       Initial  Calibration  of Multicomponent  Analytes
               (Form  VI PEST-3)
       Analyte  Resolution Summary  (Form  VI PEST-4)
       Performance Evaluation Mixture (Form VI  PEST-5)
       Individual  Standard  Mixture  A (Form VI PEST-6)
       Individual  Standard  Mixture  B (Form VI PEST-7)
       Calibration Verification Summary
               (Form  VII PEST-1)
       Calibration Verification Summary
               (Form  VII PEST-2)
       Analytical  Sequence  (Form VIII  PEST)
       Florisil Cartridge Check (Form IX I^ST-1)
       Pesticide GPC  Calibration  (Form IX PEST-2)
       Pesticide  Identification Summary  for Single
         Component Analytes (Form X PEST-1)
       Pesticide  Identification Summary  for
         Multicomponent Analytes (Form X PEST-2)
       Chromatograms  and data system printouts
         A printout of retention times and corresponding
           peak  areas  or peak heights
   d. Raw QC Data
      Blank Data
      Matrix Spike/Matrix Spike Duplicate Data
                                         FORM DC-2-3                                   OLM03.1

-------
                   ORGANICS  COMPLETE  SDG  FILE  (CSF)  INVENTORY SHEET  (Cont.)
     CASE  NO. 	 SDG  NO.  	 SDG NOS.  TO  FOLLOW
                                SAS  NO.
                                                              PAGE NOs            CHECK
                                                             FROM   TO       LAB        EPA
6.  Pesticides (Cont.)
    e.  Raw GPC Data

    f.  Raw Florisil Data

7.  Miscellaneous  Data
    Original  preparation and analysis forms or copies  of
       preparation and analysis logbook pages
    Internal  sample and  sample extract transfer chain-
       of-custody records
    Screening records
    All  instrument output,  including strip charts  from
       screening activities (describe or list)
8. EPA  Shipping/Receiving  Documents
   Airbills  (i o.  of  shipments  	)
   Chain-of-Custody  Records
   Sample Tags
   Sample Log-In  Sheet  (Lab  &  DC1)
   Miscellaneous  Shipping/Receiving  Records
       (describe or list)
9.  Internal Lab Sample Transfer Records and Tracking
   Sheets  (describe or list)
10.  Other Records  (describe  or  list)

    Telephone  Communication  Log
                                          FORM DC-2-4                                   OLM03

-------
                   ORGANICS  COMPLETE  SDG FILE  (CSF)  INVENTORY  SHEET (Cont.)
     CASE NO.
                            SDG NO.
                                                  SDG NOS.  TO FOLLOW
                                SAS  NO.
11.  Comments•
Completed by:
(CLP Lab)

Verified by:
(CLP Lab)
(Signature)
(Printed Name/Title)
(Date)
(Signature)
(Printed Name/Title)
(Date)
Audited by:
(EPA)
(Signature)
(Printed Name/Title)
(Date)
                                          FORM DC-2-5
                                                                                         OLM03.0

-------
                                   EXHIBIT C
                           TARGET COMPOUND LIST AND
                     CONTRACT REQUIRED QUANTITATION LIMITS
NOTE:  Specific quantitation limits are highly matrix-dependent.  The
       quantitation limits listed herein are provided for guidance and may not
       always be achievable.

       All CRQLs are rounded to two significant figures.

       The CRQL values listed on the following pages are based on the analysis
       of samples according to the specifications given in Exhibit D.

       For soil samples, the moisture content of the samples must be used to
       adjust the CRQL values appropriately.
                                      C-l                               OLM03.0

-------
  Exhibit C - Target Compound List and Contract Required Quantitation Limits

                               Table  of  Contents

Section                                                                   Page

1.0   VOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
      LIMITS  	    3

2.0   SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED
      QUANTITATION LIMITS 	    4
3.0   PESTICIDES/AROCLORS TARGET COMPOUND LIST AND CONTRACT REQUIRED
      QUANTITATION LIMITS 	
                                      C-2                               OLM03.0

-------
                                                        Exhibit C — Section 1
                                                               Volatiles  (VOA)


1.0   VOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION LIMITS
Quantitation Limits

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.

11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.

24.
25.
26.
27.
28.

29.
30.
31.
32.
33.
Volatiles
Chloromethane
Bromomethane
Vinyl Chloride
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide
1, 1-Dichloroethene
1 , 1-Dichloroethane
1 , 2-Dichloroethene
(total)
Chloroform
1 , 2-Dichloroethane
2-Butanone
1,1, 1-Trichloroethane
Carbon Tetrachloride
Bromodichloromethane
1, 2-Dichloropropane
cis-1, 3-Dichloropropene
Trichloroethene
Dibromochloromethane
1,1, 2-Trichloroethane
Benzene
trans-1,3-
Dichloropropene
Bromoform
4-Methy 1-2 -pent anone
2-Hexanone
Tetrachloroethene
1,1,2,2-
Tetrachloroethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Xylenes (total)
CAS Number
74-87-3
74-83-9
75-01-4
75-00-3
75-09-2
67-64-1
75-15-0
75-35-4
75-34-3
540-59-0

67-66-3
107-06-2
78-93-3
71-55-6
56-23-5
75-27-4
78-87-5
10061-01-5
79-01-6
124-48-1
79-00-5
71-43-2
10061-02-6

75-25-2
108-10-1
591-78-6
127-18-4
79-34-5

108-88-3
108-90-7
100-41-4
100-42-5
1330-20-7
Water
ug/L
10
10
10
10
10
10
10
10
10
10

10
10
10
10
10
10
10
10
10
10
10
10
10

10
10
10
10
10

10
10
10
10
10
Lew
Soil
ug/Kg
10
10
10
10
10
10
10
10
10
10

10
10
10
10
10
10
10
10
10
10
10
10
10

10
10
10
10
10

10
10
10
10
10
Med.
Soil
ug/Kg
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200

1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200
1200

1200
?200
1200
1200
1200

1200
1200
1200
1200
1200
On
Column
(ng)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)

(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)
(50)

(50)
(50)
(50)
(50)
(50)

(50)
(50)
(50)
(50)
(50)
                                     C-3
                                                                       OLM03.0

-------
Exhibit C — Section 2
Semivolatiles (SVOA)
2.0   SEMIVOLATILES TARGET COMPOUND LIST AND CONTRACT REQUIRED QUANTITATION
      LIMITS
Quantitation Limits
Water
Low
Soil
Semivolatiles CAS Number ug/L ug/Kg
34.
35.

36.
37.
38.
39.
40.
41.

42.
43.

44.
45.
46.
47.
48.
49.

50.
51.

52.
53.
54.
55.

56.
57.

58.
59.
60.
61.
Phenol
bis- ( 2-Chloroethyl )
ether
2-Chlorophenol
1 , 3-Dichlorobenzene
1 , 4-Dichlorobenzene
1 , 2-Dichlorobenzene
2-Methylphenol
2,2'-oxybis (1-
Chloropropane ) 1
4-Methylphenol
N-Nitroso-di-n-
propylaraine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4-Dimethylphenol
bis ( 2-Chloroethoxy )
methane
2 , 4-Dichlorophenol
1,2, 4-Trichloro-
benzene .
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-
methylphenol
2-Methylnaphthalene
Hexachlorocyclo-
pentadiene
2 , 4, 6-Trichlorophenol
2,4, 5-Tr ichlorophenol
2-Chloronaphthalene
2-Nitroaniline
108-95-2
111-44-4

95-57-8
541-73-1
106-46-7
95-50-1
95-48-7
108-60-1

106-44-5
621-64-7

67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
111-91-1

120-83-2
120-82-1

91-20-3
106-47-8
87-68-3
59-50-7

91-57-6
77-47-4

88-06-2
95-95-4
91-58-7
88-74-4
10
10

10
10
10
10
10
10

10
10

10
10
10
10
10
10

10
10

10
10
10
10

10
10

10
25
10
25
330
330

330
330
330
330
330
330

330
330

330
330
330
330
330
330

330
330

330
330
330
330

330
330

330
830
330
830
Med.
Soil
ug/Kg
10000
10000

10000
10000
10000
10000
10000
10000

10000
10000

10000
10000
10000
10000
10000
10000

10000
10000

10000
10000
10000
10000

10000
10000

10000
25000
10000
25000
On
Column
(ng)
(20)
(20)

(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)

(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)

(20)
(20)
(20)
(20)

(20)
(20)

(20)
(50)
(20)
(50)
     1Previously known by the name bis(2-Chloroisopropyl) ether.

                                      C-4
OLM03.0

-------
                  Exhibit C — Section 2
                    Semivolatiles (SVGA)
Quantitation Limits

62.
63.
64.
65.
66.
67.
68.
69.
70.
71.
72.

73.
74.
75.

76.

77.

78.
79.
80.
81.
82.
83.
84.
85.
86.
87.

88.
89.
90.

91.
92.
93.
Semivolatiles
Dimethylphthalate
Acenaphthylene
2 , 6-Dinitrotoluene
3-Nitroaniline
Acenaphthene
2 , 4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 , 4-Dinitrotoluene
Diethylphthalate
4-Chlorophenyl-
phenyl ether
Fluorene
4-Nitroaniline
4, 6-Dinitro-2-
methylphenol
N-Nitroso-
diphenylamine
4-Bromophenyl-
phenylether
Hexachlorobenzene
Pentachlorophenol
Phenanthrene
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3'-
Dichlorobenzidine
Benzo ( a ) anthracene
Chrysene
bis ( 2-Ethylhexyl )
phthalate
Di-n-octylphthalate
Benzo(b) f luoranthene
Benzo(k) f luoranthene
CAS Number
131-11-3
208-96-8
606-20-2
99-09-2
83-32-9
51-28-5
100-02-7
132-64-9
121-14-2
84-66-2
7005-72-3

86-73-7
100-01-6
534-52-1

86-30-6

101-55-3

118-74-1
87-86-5
85-01-8
120-12-7
86-74-8
84-74-2
206-44-0
129-00-0
85-68-7
91-94-1

56-55-3
218-01-9
117-81-7

117-84-0
205-99-2
207-08-9
Water
ug/L
10
10
10
25
10
25
25
10
10
10
10

10
25
25

10

10

10
25
10
10
10
10
10
10
10
10

10
10
10

10
10
10
Low
Soil
ug/Kg
330
330
330
830
330
830
830
330
330
330
330

330
830
830

330

330

330
830
330
330
330
330
330
330
330
330

330
330
330

330
330
330
Med.
Soil
ug/Kg
10000
10000
10000
25000
10000
25000
25000
10000
10000
10000
10000

10000
25000
25000

10000

10000

10000
25000
10000
10000
10000
10000
10000
10000
10000
10000

10000
10000
10000

10000
10000
10000
On
Column
(ng)
(20)
(20)
(20)
(50)
(20)
(50)
(50)
(20)
(20)
(20)
(20)

(20)
(50)
(50)

(20)

(20)

(20)
(50)
(20)
(20)
(20)
(20)
(20)
(20)
(20)
(20)

(20)
(20)
(20)

(20)
(20)
(20)
C-5
                                 OLM03.0

-------
Exhibit C — Section 2
Semivolatiles (SVGA)
Quantitation Limits

94.
95.
96.
97.
Semivolatiles
Benzo ( a ) pyrene
Indeno(l,2,3-cd)-
pyrene
Dibenzo(a,h)-
anthracene
Benzo ( g , h , i ) pery lene
CAS Number
50-32-8
193-39-5
53-70-3
191-24-2
Water
ug/L
10
10
10
10
Low
Soil
ug/Kg
330
330
330
330
Med.
Soil
ug/Kg
10000
10000
10000
10000
On
Column
(ng)
(20)
(20)
(20)
(20)
                                     C-6
OLM03.0

-------
                                                        Exhibit C — Section 3
                                                Pesticides/Aroclors (PEST/ARO)
3.0   PESTICIDES/AROCLORS TARGET COMPOUND LIST AND CONTRACT REQUIRED
      QUANTITATION LIMITS2'3
                                                  Quantitation Limits
Pesticides/Aroclors
98.
99.
100.
101.
102.
103.
104.
105.
106.
107.
108.
109.
110.
111.
112 .
113.
114.
115.
116.
117.
118.
119.
120.
121.
122.
123.
124.
125.
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4, 4 '-DDE
Endrin
Endosulfan II
4, 4 '-ODD
Endosulfan sulfate
4, 4 '-DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260

CAS Number
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
309-00-2
111024-57-3
959-98-8
60-57-1
72-55-9
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
72-43-5
53494-70-5
7421-93-4
5103-71-9
5103-74-2
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
Water
ug/L
0.050
0.050
0.050
0.050
0.050
0.050
0.050
0.050
0.10
0.10
0.10
0.10
0.10
0,10
0.10
0.50
0.10
0.10
0.050
O.C50
5.0
1.0
2.0
1.0
1.0
1.0
1.0
1.0
Soil
ug/Kg
1.7
1.7
1.7
1.7
1.7
1.7
1.7
1.7
3.3
3.3
3.3
3.3
3.3
3.3
3.3
17
3.3
3.3
1.7
1.7
170
33
67
33
33
33
33
33
On Column
(pg)
5
5
5
5
5
5
5
5
10
10
10
10
10
10
10
50
10
10
5
5
500
100
200
100
100
100
100
100
     2There is  no differentiation between the preparation of low and medium soil
samples in this method for the analysis of pesticides/Aroclors.

     3The lower reporting limit for pesticide instrument blanks shall be one-half
the CRQL values for water samples.

     ^Only the  exo-epoxy  isomer (isomer  B)  of heptachlor  epoxide is reported on
the data reporting forms (Exhibit B).
                                      C-7
                                                                       OLM03.0

-------
     EXHIBIT D
ANALYTICAL METHODS
   FOR VOLATILES
     D-l/VOA                             OLM03.0

-------
                 Exhibit D - Analytical Methods for Volatiles

                               Table  of  Contents


Section                                                                   Page
1.0   SCOPE AND APPLICATION 	    4

2.0   SUMMARY OF METHOD	    5
      2.1   Water	    5
      2.2   Low Soil	    5
      2.3   Medium Soil	    5

3.0   DEFINITIONS 	    5

4.0   INTERFERENCES 	    6

5.0   SAFETY	    6

6.0   EQUIPMENT AND SUPPLIES   	    7

7.0   REAGENTS AND STANDARDS	   13
      7.1   Reagents	   13
      7.2   Standards	   13
      7.3   Storage of Standard Solutions 	   16

8.0   SAMPLE COLLECTION, PRESERVATION AND STORAGE 	   18
      8.1   Sample Collection and Preservation  	   18
      8.2   Procedure for Sample Storage  	   18
      8.3   Contract Required Holding Times 	   18

9.0   CALIBRATION AND STANDARDIZATION 	   19
      9.1   Instrument Operating Conditions 	   19
      9.2   GC/MS Calibration  (Tuning) and Ion Abundance   	   21
      9.3   Initial Calibration 	   22
      9.4   Continuing Calibration   	   26

10.0  PROCEDURE	   29
      10.1  Sample Preparation  	   29
      10.2  pH Determination (Water Samples)  	   35
      10.3  Percent Moisture Determination  	   36

11.0  DATA ANALYSIS AND CALCULATIONS	   37
      11.1  Qualitative Identification  	   37
      11.2  Calculations	   39
      11.3  Technical Acceptance Criteria for Sample Analysis  	   44
      11.4  Corrective Action  for Sample Analysis 	   45

12.0  QUALITY CONTROL	   48
      12.1  Blank Analyses	   48
      12.2  Matrix Spike/Matrix Spike Duplicate (MS/MSD)   	   51

13.0  METHOD PERFORMANCE  	   55

14.0  POLLUTION PREVENTION  	   55

                                    D-2/VOA                             OLM03.0

-------
15.0  WASTE MANAGEMENT   	    55

16.0  REFERENCES	    55

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	    56

APPENDIX A - SCREENING OF HEXADECANE EXTRACTS FOR VOLATILES  	    62
      1.0   SCOPE AND APPLICATION	    62
      2.0   SUMMARY OF METHODS	    62
      3.0   INTERFERENCES	    63
      4.0   SAFETY	    63
      5.0   EQUIPMENT AND SUPPLIES	    63
      6.0   REAGENTS AND STANDARDS	    64
      7.0   QUALITY CONTROL	    65
      8.0   CALIBRATION AND STANDARDIZATION	    66
      9.0   PROCEDURE	    66
                                    D-3/VOA                             OLM03.0

-------
Exhibit D Volatiles — Section 1
Scope and Application
1.0   SCOPE AND APPLICATION

1.1   The analytical method that follows is designed to analyze water,
      sediment and soil from hazardous waste sites for the organic compounds
      on the Target Compounds List (TCL, see Exhibit C).   The method is based
      on EPA Method 624 (Purgeables).

1.2   The method includes sample preparation, a hexadecane screening procedure
      on a GC/FID to determine the approximate concentration of organic
      constituents in the sample (Appendix A), and the actual analysis which
      is based on a purge and trap gas chromatograph/mass spectrometer (GC/MS)
      method.

1.3   Problems have been associated with the following compounds analyzed by
      this method:

      •   Chloromethane, vinyl chloride, bromomet.iane, and chloroethane can
          display peak broadening if the compounds are not delivered to the GC
          column in a tight band.

      •   Acetone, hexanone, 2-butanone, and 4-methyl-2-pentanone have poor
          purge efficiencies.

      •   1,1,1-trichloroethane and all the dichloroethanes can
          dehydrohalogenate during storage or analysis.

      •   Chloromethane can be lost if the purge flow is too fast.

      •   Bromoform is one of the compounds most likely to be adversely
          affected by cold spots and/or active sites in the transfer lines.
          Response of its quantitation ion (m/z 173) is directly affected by
          the tuning of the GC/MS to meet the instrument performance criteria
          for 4-bromofluorobenzene (BFB) at ions m/z 174/176.  Increasing the
          m/z 174/176 ratio may improve bromoform response.
                                    D-4/VOA                            OLM03.0

-------
                                         Exhibit D Volatiles — Sections 2 & 3
                                                 Summary of Method/Definitions

2.0   SUMMARY OF METHOD

2.1   Water

      An inert gas is bubbled through a 5 mL sample contained in a
      specifically designed purging chamber at ambient temperature.  The
      purgeable compounds are efficiently transferred from the aqueous phase
      to the vapor phase.  The vapor is swept through a sorbent column where
      the purgeables are trapped.  After purging is completed, the sorbent
      column is heated and backflushed with the inert gas to desorb the
      purgeable compounds onto a gas chromatographic column.  The gas
      chromatograph is temperature-programmed to separate the purgeable
      compounds which are then detected with a mass spectrometer.

2.2   Low Soil

      An inert gas is bubbled through a mixture of reagent water and 5 g of
      sample contained in a specifically designed purging chamber that is held
      at an elevated temperature.  The purgeable compounds are efficiently
      transferred from the aqueous phase to the vapor phase.  The vapor is
      swept through a sorbent column where the purgeables are trapped.  After
      purging is completed, the sorbent column is heated and backflushed with
      the inert gas to desorb the purgeable compounds onto a gas
      chromatographic column.  The gas chromatograph is temperature-programmed
      to separate the purgeable compounds which are then detected with a mass
      spectrometer.

2 . 3   Medium So 1

      A measured amount of soil is extracted with methanol.  A portion of the
      methanol extract is diluted to 5 mL with reagent water.  An inert gas is
      bubbled through this solution in a specifically designed purging chamber
      at ambient temperature.  The purgeable compounds are effectively
      transferred from the aqueous phase to the vapor phase.  The vapor is
      swept through a sorbent column where the purgeables are trapped.  After
      purging is completed, the sorbent column is heated and backflushed with
      the inert gas to desorb the purgeable compounds onto a gas
      chromatographic column.  The gas chromatograph is temperature-programmed
      to separate the purgeable compounds which are then detected with a mass
      spectrometer.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                   D-5/VOA                             OLM03.0

-------
Exhibit D Volatiles — Sections 4 &
Interferences/Safety
4.0   INTERFERENCES

4.1   Method interference may be caused by impurities in the purge gas,
      organic compounds out-gassing from the plumbing ahead of the trap, and
      solvent vapors in the laboratory.  The analytical system must be
      demonstrated to be free from contamination under the conditions of the
      analysis by running laboratory method and instrument blanks- as described
      in Section 12.  The use of non-Polytetrafluoroethylene (PTFE) tubing,
      non-PTFE thread sealants, or flow controllers with rubber components in
      the purging device should be avoided.

4.2   Samples can be contaminated by diffusion of volatile organics
      (particularly fluorocarbons and methylene chloride) through the septum
      seal into the sample during storage and handling.

4.3   Contamination by carryover can occur whenever high level and low level
      samples are sequentially analyzed.  To reduce carryover, the purging
      device and sampling syringe must be rinsed with reagent water between
      sample analyses.  For samples containing large amounts of water-soluble
      materials, suspended solids, high-boiling compounds, or high purgeable
      levels, it may be necessary to wash out the purging device with a
      detergent solution between analyses, rinse it with distilled water, and
      then dry it in an oven at 105 °C.  The trap and other parts of the
      system are also subjected to contamination; therefore, frequent bakeout
      and purging of the entire system may be required.

4.4   The laboratory where volatile analysis is performed should be completely
      free of solvents.

5.0   SAFETY

5.1   The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely defined; however, each chemical compound should be
      treated as a potential health hazard.  From this viewpoint, exposure to
      these chemicals must be reduced to the lowest possible level by whatever
      means available.  The laboratory xs responsible for maintaining a
      current awareness file of OSHA regulations regarding the safe handling
      of the chemicals specified in this method.  A reference file of material
      data hand]ing sheets should also be made available to all personnel
      involved in the chemical analysis.

5.2   The following analytes covered by this method have been tentatively
      classified as known or suspected, human or mammalian carcinogens:
      benzene, carbon tetrachloride, chloroform and vinyl chloride.  Primary
      standards of these toxic compounds should be prepared in a hood.  A
      NIOSH/MESA approved toxic gas respirator should be worn when the analyst
      handles high concentrations of these toxic compounds.
                                    D-6/VOA                            OLM03.0

-------
                                              Exhibit D Volatiles — Section 6
                                                        Equipment and Supplies
6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      buc demonstration of equivalent performance meeting the requirements of
      this SOW is the responsibility of the Contractor.  The Contractor shall
      document any use of alternate equipment or supplies in the SDG
      Narrative.

6.1   Glassware

6.1.1    Syringes - 5 mL, gas-tight with shut-off valve.  Micro syringes - 25
         /iL and larger, 0.006 inch ID needle.

6.1.2    Syringe Valve - two-way, with Luer ends (three each), if applicable
         to the purging device.

6.1.3    Pasteur Pipets - disposable.

6.1.4    Vials and Caps - 2 mL for GC.

6.1.5    Volumetric Flasks.

6.1.6    Bottle - 15 mL, screw-cap, with Teflon cap liner.

6.2   pH Paper - wide range

6.3   Balances - analytical,  capable of accurately weighing + 0.0001 g, and a
      top-loading balance capable of weighing 100 g ±0.01 g.   The balances
      must be calibrated with class S weights or known reference weights once
      per each 12-hour work shift.  The balances must be calibrated with class
      S weights at a minimum of once per month.  The balances must also be
      annually checked by a certified technician.

6.4   Purge and Trap Device - consists of three separate pieces of equipment:
      the sample purge chamber, trap, and the desorber.  Several complete
      devices are now commercially available.

6.4.1    The sample purge chamber must be designed to accept 5 mL samples with
         a water column at least 3 cm deep.  The gaseous head space between
         the water column and the trap must have a total volume of less than
         15 mL.  The purge gas must pass through the water column as finely
         divided bubbles, each with a diameter of less than 3 mm at the
         origin.  The purge gas must be introduced no more than 5 mm from the
         base of the water column.

6.4.2    The trap must be at least 25 cm long and have an inside diameter of
         at least 0.105 inch.  The trap must be packed to contain the
         following minimum lengths of absorbents: (starting from inlet) 0.5 cm
         silanized glass wool, 1 cm methyl silicone, 15 cm of 2,6-diphenylene
         oxide polymer (Tenax-GC, 60/80 mesh),  8 cm of silica gel (Davison
         Chemical, 35/60 mesh, grade 15 or equivalent) and 0.5 cm silanized
         glass wool.   A description of the trap used for analysis shall be
         provided in the SDG narrative.
                                   D-7/VOA                             OLM03.0

-------
Exhibit D Volatiles — Section 6
Equipment and Supplies
6.4.3    The desorber should be capable of rapidly heating the trap to 180 °C.
         The polymer section of the trap should not be heated higher than 180
         °C and the remaining sections should not exceed 220 °C during bakeout
         mode.

6.4.4    Trap Packing

6.4.4.1     2,6-Diphenylene oxide polymer, 60/80 mesh chromatographic grade
            (Tenax GC or equivalent).

6.4.4.2     Methyl silicone packing, 3.0 percent OV-1 on Chromasorb W, 60/80
            mesh (or equivalent).

6.4.4.3     Silica gel,  35/60 mesh, Davison,  grade 15 (or equivalent).

6.4.4.4     Alternate sorbent traps may be used if:

            •   The trap packing materials do not introduce contaminants which
                interfere with identification and quantitation of the
                compounds listed in Exhibit C (Volatiles).

            •   The analytical results generated using the trap meet the
                initial and continuing calibration technical acceptance
                criteria listed in the SOW and the CRQLs listed in Exhibit C
                (Volatiles).

            •   The trap can accept up to 1000 ng of each compound listed in
                Exhibit C (Volatiles)  without becoming overloaded.

6.4.4.4.1      The alternate trap must be designed to optimize performance.
               Follow manufacturer's instructions for the use of its product.
               Before use of any trap, other than the one specified in 6.4.2,
               the Contractor must first meet the criteria listed in 6.4.4.4.
               Once this has been demonstrated, the Contractor must document
               its use in each SDG Narrative by specifying the trap
               composition (packing material/brand name, amount of packing
               material).  Other sorbent traps include, but are not limited
               to, Tenax/Silica Gel/Carbon Trap from EPA Method 524.2, Tenax -
               GC/Graphpac-D Trap  (Alltech) or equivalent, and Vocarb 4000
               Trap (Supelco) or equivalent.

6.4.4.4.2      The Contractor must maintain documentation that the alternate
               trap meets the criteria listed in 6.4.4.4.  The minimum
               documentation requirements are as follows:

6.4.4.4.2.1       Manufacturer provided information concerning the performance
                  characteristics of the trap.

6.4.4.4.2.2       Reconstructed  ion chromatograms and data system reports
                  generated on the Contractor's GC/MS used for CLP analyses:

                  •     From instrument blank analyses which demonstrate that
                        there are no contaminants which  interfere with the
                        volatile analysis when using the alternate trap;

                                    D-8/VOA                            OLM03.1

-------
                                              Exhibit D Volatiles — Section 6
                                                        Equipment and Supplies


                  •     From initial and continuing calibration standards
                        analyzed using the trap specified in Section 6.4.4.

6.4.4.4.2.3       Based on Contractor generated data described above, the
                  Contractor must complete a written comparison/review, which
                  has been signed by the Laboratory Manager, certifying that:

                  •     The alternate trap performance meets the technical
                        acceptance criteria listed in Sections 9.3.5 and
                        9.4.5;

                  •     The low point initial calibration standard analysis
                        has adequate sensitivity to meet the volatile CRQLs;

                  •     The high point initial calibration standard analysis
                        was not overloaded;

                  •     The alternate trap materials do not introduce
                        contaminants which interfere with the identification
                        and/or quantitation of the compounds listed in Exhibit
                        C (Volatiles).

6.4.4.4.2.4       The documentation must be made available to the Agency
                  during on-site laboratory evaluations or sent to the Agency
                  upon request of the Technical Project Officer or the
                  Administrative Project Officer.

6.4.5    The purge and trap apparatus may be assembled as a separate unit or
         be an integral unit coupled with a gas chromatograph.

6.5   A heater or heated bath capable of maintaining the purge chamber at 40
      °C ± 1 °C is to be used for low level soil/sediment analysis, but not
      for water or medium level soil/sediment analyses.

6.6   Gas Chromatography/Mass Spectrometer (GC/MS) System

6.6.1    Gas Chromatograph - the gas chromatograph (GC) system must be capable
         of temperature programming and have a flow controller that maintains
         a constant column flow rate throughout desorption and temperature
         program operations.  The system must include or be interfaced to a
         purge and trap system as specified in Section 6.4 and have all
         required accessories including syringes,  analytical columns, and
         gases.  All GC carrier gas lines must be constructed from stainless
         steel or copper tubing.  Non-PTFE thread sealants, or flow
         controllers with rubber components are not to be used.

6.6.2    Gas Chromatography Columns
         A description of the column used for analysis shall be provided in
         the SDG narrative.

6.6.2.1     Packed columns - 6 ft.  long x 0.1 inch ID glass, packed with 1.0
            percent SP-1000 on Carbopack B (60/80) mesh or equivalent.
                                   D-9/VOA                             OLM03.0

-------
Exhibit  D  Volatiles  —  Section  6
Equipment  and Supplies



6.6.2.2          Capillary Columns

6.6.2.2.1            Minimum length 30 m x  0.53 mm ID VOCOL (Supelco)  or  equivalent  fused  silica widebore
                     capillary column with  3 pm film thickness.

6.6.2.2.2            Minimum length 30 m x  0.53 mm ID DB-624 (J  & U Scientific)  or equivalent  fused silica
                     widebore capillary column with 3 ^m film thickness.

6.6.2.2.3            Minimum length 30 m x  0.53 mm ID AT-624 (Alltech) or equivalent fused silica widebore
                     capillary column with  3 /jn film thickness.

6.6.2.2.4            Minimum length 30 m x  0.53 mm ID HP-624 (Hewlett-Packard)  or equivalent fused silica
                     widebore capillary column with 3 (jn film thickness.

6.6.2.2.5            Minimum length 30 m x  0.53 mm ID RTx-624 (Restek) or equivalent fused silica widebore
                     capillary column with  3 /
-------
                                              Exhibit D Volatiles — Section 6
                                                        Equipment and Supplies
                •    From  initial  and  continuing calibration standards analyzed
                    using the column.

6.6.2.5.3       Based on  the Contractor-generated data described above, the
                Contractor shall  complete a written review, signed by the
                Laboratory Manager, certifying that:

                •    The column performance meets the technical acceptance
                    criteria in Sections 9.3.5 and 9.4.5;

                •    The low point initial calibration standard analysis has
                    adequate sensitivity to meet the volatile CRQLs;

                •    The high point initial calibration standard analysis was
                    not overloaded;

                •    The column does not introduce contaminants which interfere
                    with  the identification and/or quantitation of compounds
                    listed in Exhibit C (Volatiles).

6.6.2.5.4       The  documentation must be made available to the Agency during
                on-site laboratory evaluations or sent to the Agency upon
                request of the Technical Project Officer or Administrative
                Project Officer.

6.6.3    Mass Spectrometer - must be capable of scanning from 35 to 300 amu
         every  1 second  or less to every 2 seconds or less utilizing 70 volts
         (nominal)  electron energy in the electron impact ionization mode, and
         producing  a mass spectrum which meets all the instrument performance
         acceptance criteria when 50 ng of BFB is injected through the gas
         chromatograph inlet.   The instrument conditions required for the
         acquisition of  the BFB mass spectrum are given in Section 9.

6.6.3.1     NOTE: The MS scan rate should allow acquisition of at least five
            spectra while a sample compound elutes from the GC.   The purge and
            trap GC/MS system must be in a room whose atmosphere is
            demonstrated to be free of all potential contaminants which will
            interfere with the analysis.   The instrument must be v°nted to the
            outside of the facility or to a trapping system which prevents the
            release of contaminants into the instrument room.

6.6.4    GC/MS  interface - any gas chromatograph to mass spectrometer
         interface that  gives acceptable calibration points at 50 ng or less
         per injection for each of the parameters of interest, and achieves
         all acceptance  criteria, may be used.  Gas chromatograph to mass
         spectrometer interfaces constructed of all-glass or glass-lined
         materials are recommended.   Glass can be deactivated by silanizing
         with dichlorodimethylsilane.

6.6.5    Data system - a computer system must be interfaced to the mass
         spectrometer that allows the continuous acquisition and storage, on
         machine-readable media,  of all mass spectra obtained throughout the
         duration of the chromatographic program.  The computer must have

                                   D-ll/VOA                            OLM03.1

-------
Exhibit D Volatiles — Section 6
Equipment and Supplies
         software that allows searching any GC/MS data file for ions of a
         specified mass and plotting such ion abundances versus time or scan
         number.  This type of plot is defined as an Extracted Ion Current
         Profile (EICP).  Software must also be available that allows
         integrating the abundance in any EICP between specified time or scan
         number limits.  Also, for the non-target compounds, software must be
         available that allows for the comparison of sample spectra against
         reference library spectra.  The NIST/EPA/NIH (May 1992 release or
         later) and/or Wiley  (1991 release or later), or equivalent mass
         spectral library shall be used as the reference library.  The data
         system must be capable of flagging all data files that have been
         edited manually by laboratory personnel.

6.6.6    Magnetic tape storage device - must be capable of recording data and
         must be suitable for long-term, off-line storage.
                                   D-12/VOA                             OLM03.0

-------
                                              Exhibit D Volatiles — Section 7
                                                        Reagents and Standards
7.0   REAGENTS AND STANDARDS

7.1   Reagents

7.1.1    Reagent water - defined as water in which an interferent is not
         observed at or above the CRQL of the analytes of interest.  Reagent
         water may be generated by passing tap water through a carbon filter
         bed containing about 453 g (1 Ib) of activated carbon (Calgon Corp.,
         Filtrasorb-300 or equivalent).

7.1.1.1     A water purification system (Millipore Super-Q or equivalent) may
            be used to generate reagent water.

7.1.1.2     Reagent water may also be prepared by boiling water for 15
            minutes.  Subsequently, while maintaining the temperature at
            90 °C, bubble a contaminant-free inert gas through the water for
            one hour.  While still hot, transfer the water to a narrow-mouth
            screw-cap bottle and seal with a Teflon-lined septum and cap.

7.1.2    Methanol - pesticide quality or equivalent

7.2   Standards

7.2.1    Introduction

         The Contractor must provide all standards to be used with this
         contract.  These standards may be used only after they have been
         certified according to the procedure in Exhibit E.  The Contractor
         must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

7.2.2    Stock Standard Solutions

7.2.2.1     Stock standard solutions may be purchased or may be prepared in
            methanol from pure standard materials.

7.2.2.2     Prepare stock standard solutions by placing about 9.8 mL of
            methanol into a 10 mL ground-glass stoppered volumetric flask.
            Allow the flask to stand, unstoppered, for about 10 minutes, or
            until all alcohol-wetted surfaces have dried.  Weigh the flask to
            the nearest 0.1 mg.

7.2.2.3     Add the assayed reference material as described below.

7.2.2.3.1      If the compound is a liquid, using a 100 /uL syringe,
               immediately add two or more drops of assayed reference material
               to the flask, then reweigh.  The liquid must fall directly into
               the alcohol without contacting the neck of the flask.

7.2.2.3.2      If the compound is a gas at room temperature, fill a 5 mL
               valved gas-tight syringe with the reference standard to the
               5 mL mark.  Lower the needle to 5 mm above the methanol
               meniscus.  Slowly introduce the reference standard above the

                                   D-13/VOA                            OLM03.0

-------
 Exhibit  D  Volatiles  —  Section  7
 Reagents and  Standards
                surface of the  liquid.  The gas will rapidly dissolve  in the
                methanol.

 7.2.2.3.3       The procedure in Section  7.2.2.3.2 may also be accomplished by
                using a lecture bottle equipped with a Hamilton Lecture Bottle
                Septum (#86600).  Attach  Teflon tubing to the side-arm relief
                valve and direct a gentle stream of the reference standard into
                the methanol meniscus.

 7.2.2.3.4       Reweigh, dilute to volume, stopper, then mix by inverting the
                flask several times.  For non-gaseous compounds, calculate the
                concentration in micrograms per microliter from the net gain in
                weight.  When compound purity is assayed to be 97.0 percent or
                greater, the weight may be used without correction to  calculate
                the concentration of the  stock standard.  If the compound
                purity is assayed to be less than 97.0 percent, the weight must
                be corrected when calculating the concentration of the stock
                solution.   See Exhibit E  (Analytical Standards Requirements).
                For gaseous compounds, calculate the concentration in
               micrograms per microliter, using the Ideal Gas Law, taking into
                account the temperature and pressure conditions within the
                laboratory.

 7.2.2.3.5      Prepare fresh stock standards every two months for gases or for
               reactive compounds such as styrene.  All other stock standards
                for non-gases/non-reactive purgeable compounds must be replaced
               after six  months,  or sooner if the standard has degraded or
               evaporated.

7.2.3    Secondary Dilution Standards

7.2.3.1     Using stock standard solutions,  prepare secondary dilution
            standards in  methanol that contain the compounds of interest,
            either singly or mixed together.   Secondary dilution standard
            solutions should be prepared at  concentrations that can be easily
            diluted to prepare wording standard solutions.

7.2.3.2     Prepare fresh secondary dilution standards for gases and for
            reactive compounds such as styrene every month,  or sooner, if
            standard has  degraded or evaporated.   Secondary dilution standards
            for the other purgeable compounds must be replaced after six
            months,  or sooner,  if standard has degraded or evaporated.

7.2.4    Working Standards

7.2.4.1     System Monitoring Compound (SMC)  Spiking Solution

            Prepare a system monitoring compound  spiking solution containing
            toluene-dg, 4-bromofluorobenzene  (BFB),  and  l,2-dichloroethane-d4
            in methanol at a concentration of 25  jug/mL.   Add 10 fjiL of this
            spiking solution into 5 mL of sample,  sample extract or
            calibration standard  for a concentration of 50 /*g/L.   Prepare
            fresh spiking solution weekly, or sooner if the solution has
            degraded or evaporated.

                                   D-14/VOA                             OLM03.0

-------
                                              Exhibit  D  Volatiles  —  Section  7
                                                         Reagents and  Standards
7.2.4.2     Matrix  Spiking  Solution

            Prepare a  spiking  solution  in methanol  that  contains the following
            compounds  at  a  concentration of  25 /tg/mL:  1,1-dichloroethene,
            trichloroethene, chlorobenzene,  toluene,  and benzene.   Prepare
            fresh spiking solution weekly, or sooner  if  the  solution has
            degraded or evaporated.

7.2.4.3     Internal Standard  Spiking Solution

            Prepare an internal  standard spiking  solution containing
            bromochloromethane,  chlorobenzene-ds, and  1,4-difluorobenzene in
            methanol at a concentration of 25 /xg/mL for  each internal
            standard.  Add  10  jtL of this spiking  solution into  5 mL of  sample
            or calibration  standard for a concentration  of 50 ;ug/L.   Prepare
            fresh spiking solution weekly, or sooner  if  the  solution has
            degraded or evaporated.

7.2.4.4     Instrument Performance Check Solution - 4-Bromofluorobenzene  (BFB)

            Prepare a  25  ng/^L solution of BFB in methanol.  Prepare fresh  BFB
            solution every  six months, or sooner  if the  solution has degraded
            or evaporated.  NOTE: The 25 ng/juL concentration is  used with a
            2 /iL injection  volume.  The laboratory may prepare  a 50 ng//*L
            solution of BFB if a 1 jiL injection volume is used.

7.2.4.5     Calibration Standard Solution

            Prepare a  calibration standard solution containing  all  of the
            purgeable  target compounds in methanol.   The  recommended
            concentration of the target compounds is  100  /tg/mL.   Prepare fresh
            calibration standard solutions weekly, or  sooner if  solutions have
            degraded or evaporated.

7.2.4.6     Initial  and Continuing Calibration Standards

7.2.4.6.1      Prepare five aqueous initial  calibration  standard solutions
               containing all of the purgeable target  compounds  and system
               monitoring compounds at 10, 20, 50, 100 and 200 /xg/L  levels.
               It is required that all three xylenp isomers  (o-,  m-,  and p-
               xylene) be present in the calibration  standards  at
               concentrations of each isomer equal to  that of the other target
               compounds  (i.e., 10, 20,  50,  100 and 200 /Kj/L).   Similarly,  the
               cis  and trans isomers of 1,2-dichloroethene must  both be
               present in the standards at concentrations of each isomer equal
               to that of the other target compounds.

7.2.4.6.2      Aqueous calibration standards may be prepared in  a volumetric
               flask or in  the syringe used  to inject  the standard  into the
               purging device.

7.2.4.6.2.1       Volumetric flask - add an  appropriate volume of the 100
                  p.g/rnL calibration standard solution  (Section  7.2.4.5) to  an
                  aliquot of reagent water in a volumetric flask.   Use  a

                                   D-15/VOA                             OLM03.0

-------
 Exhibit  D  Volatiles  —  Section  7
 Reagents and  Standards
                  microsyringe and rapidly inject the alcohol standard  into
                  the expanded area of the filled volumetric flask.  Remove
                  the needle as quickly as possible after injection.  Bring to
                  volume.  Mix by inverting the flask three times only.
                  Discard the contents contained in the head of the flask.

 7.2.4.6.2.2       Syringe - remove the plunger from a 5 mL "Luerlock" syringe.
                  Pour reagent water into the syringe barrel to just short of
                  overflowing.  Replace the syringe plunger and compress the
                  water.  Invert the syringe, open the syringe valve and vent
                  any residual air.  Adjust the water volume to 5 mL minus the
                  amount of calibration standard to be added.  Withdraw the
                  plunger slightly and add an appropriate volume of working
                  calibration standard through the valve bore of the syringe.
                  Close the valve and invert three times.

 7.2.4.6.2.3       The 50 jug/L aqueous calibration standard solution is the
                  continuing calibration standard.

 7.2.4.6.3      The methanol contained in each of the aqueous calibration
               standards must not exceed 1.0 percent by volume.

 7.2.5    Ampulated Standard Extracts

         Standard solutions purchased from a chemical supply house as
         ampulated extracts in glass vials may be retained for 2 years from
         the preparation date,  unless the manufacturer recommends a shorter
         time period.  Standard solutions prepared by the Contractor which are
         immediately ampulated in glass vials may be retained for 2 years from
         the preparation date.   Upon breaking the glass seal,  the expiration
         times listed in Sections 7.2.2 through 7.3 will apply.   The
         Contractor is responsible for assuring that the integrity of the
         standards has not degraded (see Section 7.3.5).

 7.3   Storage of Standard Solutions

 7.3.1    Store the stock standards in Teflon-sealed screw-cap bottles with
         zero headspace at -10 °C to -20 °C,  and protect the standards from
         light.  Once one of the bottles containing the stock standard
         solution has been opened,  it may be used for no longer than one week.

7.3.2    Store secondary dilution standards in Teflon-sealed screw-cap bottles
         with minimal headspace at -10 °C to -20 °C,  and protect the standards
         from light.   The secondary dilution standards must  be checked
         frequently for signs of degradation or evaporation,  especially just
         prior to preparing working standards from them.

7.3.3    Aqueous standards may be stored for up to 24 hours  if held in Teflon-
         sealed screw-cap vials with zero headspace at 4 °C  (±2 °C) .   Protect
         the standards from light.   If not so stored,  they must  be discarded
         after one hour unless they are set up to be purged  by an autosampler.
         When using an autosampler,  the standards may be kept for up to 12
         hours in purge tubes connected via the autosampler  to the purge and
                                   D-16/VOA                            OLM03.0

-------
                                              Exhibit D Volatiles — Section 7
                                                        Reagents and Standards
         trap device.  All other working standards may be stored at -10 °C to
         -20 °C.

7.3.4    Purgeable standards must be stored separately from other standards.

7.3.5    The Contractor is responsible for maintaining the integrity of
         standard solutions and verifying prior to use.  This means that
         standards must be brought to room temperature prior to use, checked
         for losses,  and checked that all components have remained in the
         solution.
                                   D-17/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 8
Sample Collection, Preservation and Storage
8.0   SAMPLE COLLECTION, PRESERVATION AND STORAGE

8.1   Sample Collection and Preservation

8.1.1    Water samples may be collected in glass containers having a total
         volume of at least 40 mL with a Teflon-lined septum and an open top
         screw-cap.  Soil samples may be collected in glass containers or
         closed end tubes (e.g., brass sleeves)  in sufficient quantity to
         perform the analysis.  Headspace should be avoided.  The specific
         requirements for site sample collection are outlined by the Region.

8.1.2    For collection of water samples, the containers must be filled in
         such a manner that no air bubbles pass  through the sample as the
         container is being filled.  Seal the vial so that no air bubbles are
         entrapped in it.

8.1.3    Water samples are preserved to a pH of  2 at the time of collection.

8.1.4    All samples must be iced or refrigerated at 4 °C (±2 °C)  from the
         time of collection until analysis.

8.2   Procedure for Sample Storage

8.2.1    The samples must be protected from light and refrigerated at 4 °C  (±2
         °C)  from the time of receipt until 60 days after delivery of a
         reconciled, complete sample data package to the Agency.  After 60
         days, the samples may be disposed of in a manner that complies with
         all applicable regulations.

8.2.2    The samples must be stored in an atmosphere demonstrated to be free
         of all potential contaminants and in a  refrigerator used only for
         storage of volatile samples.

8.2.3    All volatile samples in an SDG must be  stored together in the same
         refrigerator.

8.2.4    Storage blanks shall be stored with samples until all samples are
         analyzed.

8.2.5    Samples, sample extracts and standards  must be stored separately.

8.2.6    Volatile standards must be stored separately from semivolatile and
         pesticide/Aroclor standards.

8.3   Contract Required Holding Times

      Analysis of water and soil/sediment samples must be completed within  10
      days of Validated Time of Sample Receipt  (VTSR).  As part of the
      Agency's QA program, the Agency may provide Performance Evaluation
      samples as standard extracts which the Contractor is required to prepare
      per the instructions provided by the Agency.  The contract required 10
      day holding time does not apply to Performance Evaluation Samples
      received as standard extracts.
                                   D-18/VOA                            OLM03.0

-------
                                              Exhibit D Volatiles — Section 9
                                               Calibration and Standardization
                                               Instrument Operating Conditions
9.0   CALIBRATION AND STANDARDIZATION

9.1   Instrument Operating Conditions

9.1.1    Purge and Trap

9.1.1.1     The following are the recommended purge and trap analytical
            conditions.  The conditions are recommended unless otherwise
            noted.

        Purge Conditions

        Purge Gas:                     Helium or Nitrogen
        Purge Time:                    11.0 ±0.1 minute
        Purge Flow Rate:               25-40 mL/minute
        Purge Temperature:             Ambient temperature for water
                                       or medium level soil/sediment
                                       samples (required); 40 °C low
                                       level soil/sediment samples
                                       (required)

        Desorb Conditions

        Desorb Temperature:            180 °C
        Desorb Flow Rate:              15 mL/minute
        Desorb Time:                   4.0 ±0.1 rr.inute
        Trap Reconditioning Conditions

        Reconditioning Temperature:    180 °C
        Reconditioning Time:           7.0 ± 0.1 minute (minimum).  A
                                       longer time may be required to
                                       bake contamination or water
                                       from the system.

9.1.1.2     Before initial use, condition the trap overnight at 180 °C by
            backflushing with at least 20 mL/minute flow of inert gas.  Do not
            vent the trap effluent onto the analytical column.  Prior to daily
            use, condition the trap by heating at 180 °C for 10 minutes while
            backflushing.  The trap may be vented to the analytical column
            during daily conditioning; however, the column must be run through
            the temperature program prior to the analysis of samples.

9.1.1.3     Optimize purge and trap conditions for sensitivity and to minimize
            cross-contamination between samples.  Once optimized, the same
            purge and trap conditions must be used for the analysis of all
            standards, samples, and blanks.

9.1.1.4     A moisture reduction/water management system may be used to
            improve the chromatographic performance by controlling moisture or
            water if:

                                   D-19/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 9
Calibration and Standardization
Instrument Operating Conditions
9.1.2
9.1.2.1
   •   The system does not introduce contaminants which interfere
       with identification and guantitation of compounds listed in
       Exhibit C (Volatiles),

   •   The analytical results generated when using the moisture
       reduction/water management system meet the initial and
       continuing calibration technical acceptance criteria listed in
       the SOW and the CRQLs listed in Exhibit C {Volatiles),

   •   All calibration standards and samples, including blanks and
       MS/MSDs are analyzed under the same conditions,

   •   The Contractor performs acceptably on the Performance
       Evaluation samples using this system.

Gas Chromatograph

   The following are the recommended GC analytical conditions.  The
   conditions are recommended unless otherwise noted.
  Packed columns
  Carrier Gas:
  Flow Rate:
  Initial Temperature:
  Initial Hold Time:
  Ramp Rate:
  Final Temperature:
  Final Hold Time:
  Transfer Line Temperature:
                            Helium
                            30 mL/minute
                            45 °C
                            3 minutes
                            8 C°/minute
                            220 °C
                            Until three minutes after all
                            compounds listed in Exhibit C
                            (Volatiles) elute (required)
                            250-300 °C
  Capillary Columns
  Carrier Gas:
  Flow Rate:
  Initial Temperature:
  Initial Hold Time:
  Ramp Rate:
  Final Temperature:
  Final Hold Time:
                            Helium
                            15 mL/minute
                            10 °C
                            1.0 - 5.0 (±0.1)  minutes
                            6 C°/minute
                            160 °C
                            Until three  minutes after all
                            compounds listed in Exhibit C
                            (Volatiles)  elute (required)
9.1.2.2     Optimize GC conditions for analyte separation and sensitivity.
            Once optimized, the same GC conditions must be used for the
            analysis of all standards, samples, blanks, matrix spikes and
            matrix spike duplicates.

9.1.2.3     For capillary columns, if the gaseous compounds chloromethane,
            bromomethane, vinyl chloride, and chloroethane fail to exhibit
                                   D-20/VOA
                                                              OLM03.0

-------
                                              Exhibit D Volatiles — Section 9
                                               Calibration and Standardization
                                           GC/MS Calibration and Ion Abundance
            narrow, symmetrical peak shape, are not separated from the solvent
            front, or are not resolved greater than 90.0 percent from each
            other, then a subambient oven controller must be used, and the
            initial temperature must be less than or equal to 10 °C.

9.1.3    Mass Spectrometer

         The following are the required mass spectrometer analytical
         conditions:

            Electron  Energy:    70  volts  (nominal)
            Mass  Range:         35-300  amu
            Scan  Time:          To  give at  least  5  scans  per  peak,
                                not to  exceed  2 seconds per scan
                                for capillary  column.

                                To  give at  least  5  scans  per  peak,
                                not to  exceed  3 seconds per scan
                                for packed  column.

9.2   GC/MS Calibration (Tuning) and Ion Abundance

9.2.1    Summary of GC/MS Performance Check

9.2.1.1     The GC/MS system must be tuned to meet the manufacturer's
            spe ifications, using a suitable caiibrant such as perfluoro-tri-
            n-butylamine (FC-43) or perfluorokerosene (PFK).   The mass
            calibration and resolution of the GC/MS system are verified by the
            analysis of the instrument performance check solution (Section
            7.2.4.4).

9.2.1.2     Prior to the analysis of any samples, including MS/MSDs,  blanks,
            or calibration standards, the Contractor must establish that the
            GC/MS system meets the mass spectral  ion abundance criteria for
            the instrument performance check solution containing BFB.

9.2.2    Frequency of GC/MS Performance Check

         The instrument performance check solution must be injected once at
         the beginning of each 12-hour period, during which samples or
         standards are to be analyzed.   The twelve (12)  hour time period for
         GC/MS instrument performance check (BFB),  standards calibration
         (initial or continuing calibration criteria), blank and sample
         analysis begins at the moment of injection of the BFB analysis that
         the laboratory submits as documentation  of a compliant instrument
         performance check.  The time period ends after twelve (12) hours have
         elapsed according to the system clock.
                                   D-21/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 9
Calibration and Standardization
GC/MS Calibration and Ion Abundance
9.2.3    Procedure for GC/MS Performance check

9.2.3.1     The analysis of the instrument performance check solution may be
            performed as follows:

            •   As an injection of up to 50 ng of BFB into the GC/MS.

            •   By adding 50 ng of BFB to 5 mL of reagent water and analyzing
                the resulting solution as if it were an environmental sample
                (see Section 10).

9.2.3.2     The instrument performance check solution must be analyzed alone
            without calibration standards.  NOTE: The calibration standards
            contain BFB as a system monitoring compound (SMC).

9.2.4    Technical Acceptance Criteria for GC/MS Performance Check

9.2.4.1     The mass spectrum of BFB must be acquired in the following manner.
            Three scans (the peak apex scan and the scans immediately
            preceding and following the apex)  are acquired and averaged.
            Background subtraction is required,  and must be accomplished using
            a single scan no more than 20 scans prior to the elution of BFB.
            Do not background subtract part of the BFB peak.

9.2.4.2     NOTE:  All subsequent standards, samples, MS/MSD, and blanks
            associated with a BFB analysis must use identical mass
            spectrometer instrument conditions.

9.2.4.3     The analysis of the instrument performance check solution must
            meet the ion abundance criteria given in Table 1.

9.2.5    Corrective Action for GC/MS Performance Check

9.2.5.1     If the BFB technical acceptance criteria are not met, retune the
            GC/MS system.   It may also be necessary to clean the ion source,
            clean the quadrupole rods, or take other corrective actions to
            achieve the technical acceptance criteria.

9.2.5.2     BFB technical acceptance criteria must, be met before any
            standards, samples, including MS/MSDs or required blanks are
            analyzed.  Any samples or required blanks analyzed when tuning
            technical acceptance criteria have not been met will require
            reanalysis at no additional cost to the Agency.

9.3   Initial Calibration

9.3.1    Summary of Initial Calibration

         Prior to the analysis of samples and required blanks, and after the
         instrument performance check solution criteria have been met, each
         GC/MS system must be calibrated at five concentrations
                                   D-22/VOA                            OLM03.1

-------
                                              Exhibit D Volatiles — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
         to determine  instrument sensitivity and the linearity of GC/MS
         response  for  the purgeable target compounds.

9.3.2    Frequency of  Initial Calibration

9.3.2.1     Each GC/MS system must be calibrated upon award of the contract,
            whenever the Contractor takes corrective action which may change
            or affect  the initial calibration criteria (e.g., ion source
            cleaning or repair, column replacement, etc.), or if the
            continuing calibration acceptance criteria have not been met.

9.3.2.2     If time remains in the 12-hour time period after meeting the
            technical  acceptance criteria for the initial calibration, samples
            may be analyzed.  It is not necessary to analyze a continuing
            calibration standard if the initial calibration standard that is
            the same concentration as the continuing calibration standard
            meets the  continuing calibration technical acceptance criteria.  A
            method blank is required.  Quantify all sample and quality control
            sample results, such as internal standard area response change and
            retention  time shift, against the initial calibration standard
            that is the same concentration as the continuing calibration
            standard.

9.3.3    Procedure for Initial Calibration

9.3.3.1     Assemble a purge and trap device that meets the specifications in
            Section 6.4.   Condition the device as described in Section 9.1.1.

9.3.3.2     Connect the purge and trap device to the gas chromatograph.  The
            gas chromatograph must be operated using temperature and flow rate
            parameters equivalent to those in 9.1.2.

9.3.3.3     Add 10 ^L of the internal standard solution (Section 7.2.4.3) to
            each of the five aqueous calibration standard solutions containing
            the system monitoring compounds (Section 7.2.4.6) for a
            concentration of 50 /xg/L at time of purge.   Analyze each
            calibration standard according to Section 10.

9.3.3.4     Separate initial and continuing calibrations must be performed for
            water samples and low level soil/sediment samples (unheated purge
            vs. heated purge).  Extracts of medium level soil/sediment samples
            may be analyzed using the calibrations of water samples.

9.3.4    Calculations for Initial Calibration

9.3.4.1     Calculate the relative reponse factor (RRF) for each volatile
            target and system monitoring compound using equation 1.  The
            primary characteristic ions used for quantitation are listed in
            Table 2 and Table 4.  Assign the target compounds, and system
            monitoring compound to an internal standard according to Table 3.
            If an interference prevents the use of a primary ion for a given
            internal standard, use a secondary ion listed in Table 4.  NOTE:
                                   D-23/VOA                            OLM03.1

-------
Exhibit D Volatiles — Section 9
Calibration and Standardization
Initial Calibration
            Unless otherwise stated, the area response of the primary
            characteristic ion is the quantitation ion.
                                            AC
                                      RRF = -~2L  x ±i2
            Where,

            Ax  =  Area of the characteristic ion (EICP)  for the compound to
                   be measured (see Table 2)
            Ajs  =   Area of  the characteristic  ion  (EICP)  for  the  specific
                   internal standard (see Tables 3 and 4)
            CjS  =   Concentration of the  internal  standard
            Cx  =  Concentration of  the  compound to  be measured
9.3.4.2     Calculating the relative response factor of the xylenes and the
            cis and trans isomers of 1,2-dichloroethene requires special
            attention.   On packed columns, o- and p-xylene isomers co-elute.
            On capillary columns, the m- and p-xylene isomers co-elute.
            Therefore,  when calculating the relative response factor in the
            equation above, use the area response (Ax)  and concentration (Cx)
            of the peak that represents the single isomer on the GC column
            used for analysis.

9.3.4.3     For the cis and trans isomers of 1,2-dichloroethene which may co-
            elute on packed columns but not on capillary columns, both isomers
            must be present in the standards.  If the two isomers co-elute,
            use the area of the co-eluting peak and the total concentration of
            the two isomers in the standard to determine the relative response
            factor.  If the two isomers do not co-elute, sum the areas of the
            two peaks and the concentrations of the two isomers in the
            standard to determine the relative response factor.

9.3.4.4     The mean relative response factor (RRF) must be calculated for all
            compounds.
                                   D-24/VOA                            OLM03.0

-------
                                              Exhibit D Volatiles — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
9.3.4.5     Calculate the % Relative Standard Deviation  (%RSD) of the RRF
            values over the working range of the curve.
            EQ. 2
                        *RSD = Standard Deviation x 10Q
                                      Mean
Where,

 Standard Deviation  =
                                    \
                                      1-1
                             (n-l)
            X,- = each individual value used to calculate the mean
            X  = the mean of n values
            n  = the total number of values

9.3.5    Technical Acceptance Criteria for Initial Calibration

9.3.5.1     All initial calibration standards must be analyzed at the
            concentration levels described in Section 7.2.4.6.1, and at the
            frequency described in Section 9.3.2 on a GC/MS system meeting the
            BFB technical acceptance criteria.

9.3.5.2     The relative response factor (RRF) at each calibration
            concentration for each purgeable target and system monitoring
            compound  must be greater than or equal to the compound's minimum
            acceptable response factor listed in Table 5.

9.3.5.3     The %RSD for each target or system monitoring compound listed in
            Table 5 must be less than or equal to that value listed.

9.3.5.4     Up to two compounds may fail the criteria listed in Sections
            9.3.5.2 and 9.3.5.3 and still meet the minimum response factor and
            %RSD requirements.  However, these compounds must have a minimum
            RRF greater than or equal to 0.010, and the %RSD must be less than
            or equal to 40.0 percent.

9.3.5.5.    Excluding those ions in the solvent front, the combined xylenes
            and 1,2-Dichloroethenes in the 200 /*g/L standard, no quantitation
            ion may saturate the detector.   Follow the manufacturer's
            instrument operating manual to determine how saturation is
            indicated for your instrument.

9.3.6    Corrective Action for Initial Calibration

9.3.6.1     If the initial calibration technical acceptance criteria are not
            met, inspect the system for problems.  It may be necessary to
            clean the ion source,  change the column, service the purge and
            trap device or take other corrective actions to achieve the
            technical acceptance criteria.
                                   D-25/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 9
Calibration and Standardization
Continuing Calibration
9.3.6.2     Initial calibration technical acceptance criteria must, be met
            before any samples or required blanks are analyzed.  Any samples
            including MS/MSD or required blanks analyzed when initial
            calibration technical acceptance criteria have not been met will
            require reanalysis at no additional cost to the Agency.

9.4   Continuing Calibration

9.4.1    Summary of Continuing Calibration

         Prior to the analysis of samples and required blanks and after BFB
         and initial calibration acceptance criteria have been met, each GC/MS
         system must be routinely checked by analyzing a continuing
         calibration standard containing all the purgeable target and system
         monitoring compounds to ensure that the instrument continues to meet
         the instrument sensitivity and linearity requirements of the SOW.

9.4.2    Frequency of Continuing Calibration

9.4.2.1     A check of the calibration curve must be performed once every 12
            hours (see Section 9.2.2 for the definition of the 12-hour time
            period) .  If time remains in the 12-hour time period after meeting
            the technical acceptance criteria for the initial calibration,
            samples may be analyzed.  It is not necessary to analyze a
            continuing calibration standard if the initial calibration
            sta iard that is the same concentration as the continuing
            calibration standard meets the continuing calibration technical
            acceptance criteria.  A method blank is required.  Quantify all
            sample results against the initial calibration standard that is
            the same concentration as the continuing calibration standard (50
9.4.2.2     If time does not remain in the 12-hour period beginning with the
            injection of the instrument performance check solution, a new
            injection of the instrument performance check solution must be
            made.  If the new injection meets the ion abundance criteria for
            BFB,  then a continuing calibration standard may be injected.

9.4.3    Procedure for Continuing Calibration

9.4.3.1     Set up the purge and trap GC/MS system per the requirements in
            Section 9.1.1.

9.4.3.2     Add 10 /xL of internal standard solution (Section 7.2.4.3) to the 5
            mL syringe or volumetric flask containing the continuing
            calibration standard (Section 7.2.4.6).  Analyze the continuing
            calibration standard according to Section 10.

9.4.4    Calculations for Continuing Calibration

9.4.4.1     Calculate a relative response factor (RRF) for each target and
            system monitoring compound using Equation 1.
                                   D-26/VOA                            OLM03.0

-------
                                              Exhibit D Volatiles — Section 9
                                               Calibration and Standardization
                                                        Continuing Calibration
9.4.4.2     Calculate the percent difference between the continuing
            calibration relative response factor and the most recent initial
            calibration mean relative response factor for each purgeable
            target and system monitoring compound using Equation 3.
            EQ. 3
                                      RRF. -  RRF,
                       ^Difference =  	c	 x 100
            Where,

            RRFC   =   Relative response factor  from continuing calibration
                       standard

            RRFj   =   Mean relative response factor from the  most  recent
                       initial calibration meeting technical acceptance
                       criteria

9.4.5    Technical Acceptance Criteria for Continuing Calibration

9.4.5.1     The continuing calibration standard must be analyzed at the
            frequency described in Section 9.4.2 on a GC/MS system meeting the
            BFB and initial calibration technical acceptance criteria.

9.4.5.2     The relative response factor (RRF) for each purgeable target and
            system monitoring compound listed in Table 5 must be greater than
            or equal to the compound's minimum acceptable response factor
            listed in Table 5.

9.4.5.3     The relative response factor percent difference for each purgeable
            target and system monitoring compound listed in Table 5 must be
            less than or equal to the value listed.

9.4.5.4     Up to two compounds may fail the requirements listed in Sections
            9.4.5.2 and 9.4.5.3 and still meet the minimum relative response
            factor criteria and percent difference criteria.  However, these
            compounds must have a minimum relative response factor greater
            than or equal to 0.010 and the percent difference must be within
            the inclusive range of ±40.0 percent.

9.4.5.5     Excluding those ions in the solvent front, no quantitation ion may
            saturate the detector.  Consult the manufacturer's instrument
            operating manual to determine how saturation is indicated for your
            instrument.

9.4.6    Corrective Action for Continuing Calibration

9.4.6.1     If the continuing calibration technical acceptance criteria are
            not met, recalibrate the GC/MS instrument according to Section
            9.3.3.  It may be necessary to clean the ion source, change the
                                   D-27/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 9
Calibration and Standardization
Continuing Calibration
            column or take other corrective actions to achieve the continuing
            calibration technical acceptance criteria.

9.4.6.2     Continuing calibration technical acceptance criteria must be met
            before any samples,  which include MS/MSD samples,  or required
            blanks are analyzed.  Any samples or required blanks analyzed when
            continuing calibration technical acceptance criteria have not been
            met will require reanalysis at no additional cost  to the Agency.
                                   D-28/VOA                            OLM03.0

-------
                                              Exhibit  D  Volatiles  —  Section 10
                                                                      Procedure
                                                             Sample Preparation
 10.0   PROCEDURE

 10.1   Sample  Preparation

 10.1.1   If insufficient  sample  amount  (less than  90% of the  required  amount)
          is received to perform  the  analyses, the  Contractor  shall  contact  SMO
          to apprise them  of the  problem.   SMO will contact the Region  for
          instructions.  The Region will either require that no sample  analyses
          be performed or  will  require that a reduced volume be used for the
          sample analysis.  No  other  changes in the analyses will be permitted.
          The  Contractor shall  document the Region's decision  in the SDG
          Narrative.

 10.1.2   If multiphase samples (e.g., two-phase liquid sample, oily
          sludge/sandy soil sample) are received by the Contractor,  the
          Contractor shall contact SMO to apprise them of the  type of sample
          received.  SMO will contact the Region.   If all phases of  the sample
          are  amenable to  analysis, the Region may  require the Contractor to do
          the  following:

          •    Mix the sample and  analyze an aliquot from the homogenized
              sample.

          •    Separate the phases of the sample and analyze each phase
              separately.  SMO  will provide EPA sample numbers for the
              additional phases,  if required.

          •    Separate the phases, and analyze one  or more of  the phases, but
              not all of the phases.  SMO will provide EPA sample numbers for
              the additional phases, if required.

          •    Do not analyze the  sample.

10.1.2.1    If all of the phases are not amenable  to analysis (i.e., outside
            scope), the Region may require the Contractor to  do tae following:

            •   Separate the phases and analyze the phase(s)  that is amenable
                to analysis.    SMO will provide EPA sample numbers for  the
                additional phases, if required.

            •   Do not analyze the sample.

10.1.2.2    No other changes in the analyses will be permitted.   The
            Contractor shall document the Region's decision in the SDG
            Narrative.

10.1.3   Water Samples

10.1.3.1    All water samples must be allowed to warm to ambient temperature
            before analysis.

10.1.3.2    Prior to the analysis of samples,  establish the appropriate purge
            and trap GC/MS operating conditions,  as outlined in Section 9.1,

                                   D-29/VOA                            OLM03.0

-------
Exhibit D Volatiles —  Section 10
Procedure
Sample Preparation
            analyze the  instrument performance check solution  (9.2), and
            calibrate the GC/MS system according to Sections 9.3 through
            9.4.6.

10.1.3.3    If time remains in the 12-hour period  (as described in Section
            9.3.2), samples may be analyzed without analysis of a continuing
            calibration  standard.

10.1.3.4    If time does not remain in the 12-hour period since the injection
            of the instrument performance check solution, both the instrument
            performance  check solution and the continuing calibration standard
            must be analyzed before sample analysis may begin  (see Section
            9.4.2) .

10.1.3.5    Adjust the purge gas (helium) flow rate to 25-40 mL/minute.
            Variations from this flow rate may be necessary to achieve better
            purging and  collection efficiencies for some compounds,
            particularly chloromethane and bromoform.

10.1.3.6    Remove the plunger from a 5 mL syringe and attach a closed syringe
            valve.  Open the sample or standard bottle which has been allowed
            to come to ambient temperature, and carefully pour the sample into
            the syringe barrel to just short of overflowing.  Replace the
            syringe plunger and compress the sample.  Open the syringe valve
            and vent any residual air while adjusting the sample volume to 5
            mL.  This process of taking an aliquot destroys the validity of
            the sample for future analysis so, if there is only one VOA vial,
            the analyst must fill a second syringe at this time to protect
            against possible loss of sample integrity.   This second sample is
            maintained only until such time as the analyst has determined that
            the first sample has been analyzed properly.   Filling one 5 mL
            syringe would allow the use of only one syringe.  If an analysis
            is needed from the second 5 mL syringe, it must be performed
            within 24 hours.  Care must also be taken to prevent air from
            leaking into the syrin-,3.

10.1.3.7    Add 10 /xL of the system monitoring compound spiking solution
            (Section 7.2.4.1)  and 10 ;tL of the internal standard spiking
            solution (Section 7.2.4.3) through the valve bore of the syringe,
            then close the valve.   The system monitoring compounds and
            internal standards may be mixed and added as a single spiking
            solution.  The addition of 10 /tL of the system monitoring compound
            spiking solution to 5 mL of sample is equivalent to a
            concentration of 50 /xg/L of each system monitoring compound.

10.1.3.8    Attach the syringe-syringe valve assembly to the syringe valve on
            the purging device.  Open the syringe valves and inject the sample
            into the purging chamber.

10.1.3.9    Close both valves and purge the sample for 11.0 ± 0.1 minutes at
            ambient temperature.
                                   D-30/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 10
                                                                     Procedure
                                                            Sample Preparation
10.1.3.10   At the conclusion of the purge time, attach the trap to the
            chromatograph, adjust the device to the desorb mode, and begin the
            gas chromatographic temperature program.  Concurrently, introduce
            the trapped materials to the gas chromatographic column by rapidly
            heating the trap to 180 °C while backflushing the trap with an
            inert gas between 20 and 60 mL/minute for four minutes.

10.1.3.11   While the trap is being desorbed into the gas chromatograph, empty
            the purging chamber.  Wash the chamber with a minimum of two 5 mL
            flushes of reagent water to avoid carryover of target compounds.
            For samples containing large amounts of water-soluble materials,
            suspended solids, high-boiling compounds, or high purgeable
            levels, it may be necessary to wash out the purging device with a
            detergent solution between analyses, rinse it with distilled
            water, and then dry it in an oven at 105 °C.

10.1.3.12   After desorbing the sample for four minutes,  recondition the trap
            by returning the purge and trap device to the purge mode.  Wait 15
            seconds,  then close the syringe valve on the purging device to
            begin gas flow through the trap.  The trap temperature should be
            maintained at 180 °C.   Trap temperatures up to 220 °C may be
            employed.  However, the higher temperature will shorten the useful
            life of the trap.  After approximately seven minutes, turn off the
            trap heater and open the syringe valve to stop the gas flow
            through the trap.  When cool, the trap is ready for the next
            sample.

10.1.4   Low Level Soil/Sediment Samples

10.1.4.1    The Contractor must determine whether a soil/sediment sample
            should be analyzed by the low or medium method. It is the
            responsibility of the Contractor to analyze the sample at the
            correct level.

10.1.4.2    Three approaches may be taken to determine whether the low level
            or medium level method must be followed.

            •   Assume the sample is low level and analyze a 5 g sample.

            •   Use the X factor calculated from the hexadecane screen
                (Appendix A)  to determine the appropriate method for analysis.

            •   Use an in-house laboratory screening procedure.  This
                procedure must be documented and available for review during
                on-site laboratory evaluation or when requested by the
                Technical Project Officer or Administrative Project Officer.

10.1.4.3    If the on column concentration of any TCL compound exceeds the
            initial calibration range from the analysis of 5 g sample, a
            smaller sample size must be analyzed.  However, the smallest
            sample size permitted is 0.5 g.  If smaller than 0.5 g sample size
            is needed to prevent the on column concentration of TCL compounds
                                   D-31/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 10
Procedure
Sample Preparation
             from exceeding the initial calibration range, the medium level
             method must be used.

10.1.4.4     The low level soil/sediment method is based on a heated purge of a
             soil/sediment sample mixed with reagent water containing the
             system monitoring compounds and the internal standards.  Analyze
             all MS/MSD samples, blanks and standards under the same conditions
             as the samples.

10.1.4.5     Use 5 grams of sample, or use the X Factor (Appendix A) or your
             in-house screening procedure to determine the sample size for
             purging.

10.1.4.6     Prior to the analysis of samples, establish the appropriate purge
             and trap GC/MS operating conditions, as outlined in Section 9.1,
             analyze the instrument performance check solution (9.2), and
             calibrate the GC/MS system according to Sections 9.3 through
             9.4.6.  This should be done prior to the preparation of the sample
             to avoid loss of volatiles from standards and sample.  A heated
             purge calibration curve must be prepared and used for the
             quantitation of all samples analyzed with the low level method.
             Follow the initial and daily calibration instructions (9.3.3 and
             9.4.3),  but increase the purge temperature to 40 °C.

10.1.4.7     To prepare the reagent water containing the system monitoring
             compounds and the internal standards,  remove the plunger from a 5
             mL "Luerlok" type syringe equipped with a syringe valve and fill
             until overflowing with reagent water.   Replace the plunger and
             compress the water to vent trapped air.   Adjust the volume to 5
             mL.  Add 10 jitL of the system monitoring compound spiking solution
             and 10 /xL of the internal standard solution to the syringe through
             the valve.  NOTE:  Up to 10 mL of reagent water may be added to a
             soil sample to increase purge gas/sample interaction.  All soil
             samples including MS and MSD, standards, and blanks within an SDG
            must have the same amount of reagent water added.  Do not
             increase/change the amount of system monitoring compound and
             internal standard solution added.

10.1.4.8     The sample (for volatile organics) is defined as the entire
             contents of the sample container.  Do not discard any supernatant
             liquids.   Mix the contents of the sample container with a narrow
            metal spatula.  Weigh 5 g or the amount determined in using the
             screening procedure in Appendix A or an in-house screening
            procedure into a tared purge device.  Use a top loading balance.
            Note and record the actual weight to the nearest 0.1 g.

10.1.4.9    Add the spiked reagent water to the purge device and connect the
            device to the purge and trap system.

10.1.4.10   NOTE:  Prior to the attachment of the purge device, the steps in
            Sections 10.1.4.7 and 10.1.4.9 above must be performed rapidly to
                                   D-32/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 10
                                                                     Procedure
                                                            Sample Preparation
            avoid loss of volatile organics.  These steps must be performed in
            a laboratory free of solvent fumes.

10.1.4.11   Heat the sample to 40 °C ±1 °C and purge the sample for 11.0 ±
            0.1 minutes.

10.1.4.12   Proceed with the analysis as outlined in Sections 10.1.3.9 through
            10.1.3.12.

10.1.5   Medium Level Soil/Sediment Samples

10.1.5.1    The medium level soil/sediment method is based on extracting the
            soil/sediment sample with methanol.  An aliquot of the methanol
            extract is added to reagent water containing the system monitoring
            compounds and the internal standards.  The reagent water
            containing the methanol extract is purged at ambient temperature.
            When using the screening method in Appendix A, all samples with an
            X Factor > 1.0 should be analyzed by the medium level method.

10.1.5.2    Prior to the analysis of samples, establish the appropriate purge
            and trap GC/MS operating conditions, as outlined in Section 9.1,
            analyze the instrument performance check solution (9.2), and
            calibrate the GC/MS system according to Sections 9.3 through
            9.4.6.  This should be done prior to the addition of the methanol
            extract to reagent water.  Because the methanol extract and
            reagent water mixture is purged at ambient temperature, the
            instrument performance check, initial calibration, and continuing
            calibration for water samples may be used for analyses of medium
            level soil/sediment sample extracts.

10.1.5.3    The sample (for volatile organics) is defined as the entire
            contents of the sample container.  Do not discard any supernatant
            lig_uids.  Mix the contents of the sample container with a narrow
            metal spatula.   Weigh 4 g (wet weight) into a tared 15 mL vial.
            Use a top loading balance.   Note and record the actual weight to
            the nearest 0.1 g.

10.1.5.4    Quickly add 10 mL of methanol to the vial.  Cap and shake for 2
            minutes.

10.1.5.5    NOTE: The steps in Sections 10.1.5.3 and 10.1.5.4 must be
            performed rapidly to avoid loss of volatile organics.  These steps
            must be performed in a laboratory free of solvent fumes.

10.1.5.6    Let the solution settle.  Then, using a disposable pipette,
            transfer approximately 1 mL of extract into a GC vial for storage.
            The remainder may be discarded.  The 1 mL extract may be stored in
            the dark at less than 4 °C but not greater than 6 °C prior to the
            analysis.

10.1.5.7    Table 6 can be used to determine the volume of methanol extract to
            add to the 5 mL of reagent water for analysis.  If the hexadecane
            screen procedure (Appendix A) was followed, use the estimated

                                   D-33/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 10
Procedure
Sample Preparation
            concentration (Option A)  or the X Factor (Option B)  to determine
            the appropriate volume.   Otherwise,  estimate the concentration
            range of the sample from the low level analysis or from the in-
            house screening procedure to determine the appropriate volume.

10.1.5.8    Remove the plunger from a 5 mL "Luerlok" type syringe equipped
            with a syringe valve and fill until  overflowing with reagent
            water.  Replace the plunger and compress the water to vent trapped
            air.  Adjust the volume to 4.9 mL.  Pull the plunger back to 5 mL
            to allow volume for the addition of  sample and standards.  Add 10
            /iL of system monitoring compound and 10 /xL of the internal
            standard solution.  Also add the volume of methanol extract
            determined in Section 10.1.5.7 and a volume of clean methanol to
            total 100 /xL (excluding methanol in  system monitoring/internal
            standard solutions).

10.1.5.9    Attach the syringe-syringe valve assembly to the syringe valve on
            the purge device.   Open the syringe  valve and inject the
            water/methanol sample into the purging chamber.

10.1.5.10   Proceed with the analysis as outlined in Section 10.1.3.9 through
            10.1.3.12.

10.1.6   Sample Dilutions

10.1.6.1    For medium level soil/sediment analyses, the purgeable organics
            screening procedure (Appendix A), if used, will show the
            approximate concentrations of major  sample components.  If a
            dilution of the sample was indicated, this dilution shall be made
            just prior to GC/MS analysis of the  sample.  All steps in the
            dilution procedure must be performed without delays until the
            point at which the diluted sample is in a gas tight syringe.

10.1.6.2    If the on-column concentration of any compound in any sample
            exceeds the initial calibration range, a new aliquot of that
            sample must be diluted and purged.  Guidance in performing
            dilutions and exceptions to this requirement are given in Sections
            10.1.6.3 through 10.1.6.12.

10.1.6.3    Use the results of the original analysis to determine the
            approximate dilution factor required to get the largest analyte
            peak within the initial calibration  range.

10.1.6.4    The dilution factor chosen should keep the response of the largest
            analyte peak for a target compound in the upper half of the
            initial calibration range of the instrument.

10.1.6.5    All dilutions for water samples are made in volumetric flasks
            (10 mL to 100 mL).

10.1.6.6    Select the volumetric flask that will allow for the necessary
            dilution.  Intermediate dilutions may be necessary for extremely
            large dilutions.

                                   D-34/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 10
                                                                     Procedure
                                                              pH Determination
10.1.6.7    Calculate the approximate volume of reagent water which will be
            added to the volumetric flask selected and add slightly less than
            this quantity of reagent water to the flask.

10.1.6.8    For water samples, inject the proper aliquot from the syringe
            prepared in Section 10.1.3.6 into the volumetric flask.  Aliquots
            of less than 1 mL increments are prohibited.  Dilute the flask to
            the mark with reagent water.  Cap the flask, invert, and shake
            three times.

10.1.6.9    Fill a 5 mL syringe with the diluted sample as in Section
            10.1.3.6.

10.1.6.10   If this is an intermediate dilution, use it and repeat the above
            procedure to achieve larger dilutions.

10.1.6.11   Do not submit data for more than two analyses, i.e., the original
            sample and one dilution, or, from the most concentrated dilution
            analyzed and one further dilution.

10.1.6.12   For total xylenes, where three isomers are quantified as two
            peaks, the calibration of each peak should be considered
            separately,  i.e., a diluted analysis is not required for total
            xylenes unless the concentration of the peak representing the
            single isomer exceeds 200 ^g/L (/xg/kg for so\ls/sediment) or the
            peak representing the two co-eluting  somers on the GC column
            exceeds 400  /xg/L (/xg/kg for soils/sediment) .  Similarly, if the
            cis and trans isomers of 1,2-dichloroethene co-elute, a diluted
            analysis is  not required unless the concentration of the co-
            eluting peak exceeds 400 /ug/L (/xg/kg for soils/sediment).  If the
            two isomers  do not co-elute, a diluted analysis is not required
            unless the concentration of either peak exceeds 200 /xg/L (tig/kg
            for soils/sediment).

10.2  pH Determination (Water Samples)

      Once the sample aliquots have been taken from the VOA vial, the pH of
      the water sample must be determined.  The purpose of the pH
      determination is to ensure that all VOA samples were acidified in the
      field.   Test the pH by placing one or two drops of sample on the pH
      paper (do not add  pH paper to the vial).   Record the pH of each sample,
      and report these data in the SDG Narrative, following the instructions
      in Exhibit B.  No  pH adjustment is to be performed by the Contractor.
                                   D-35/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 10
Procedure
pH and Percent Moisture Determination


10.3  Percent Moisture Determination

      Immediately after weighing the sample for analysis,  weigh 5-10 g of the
      soil/sediment into a tared crucible.   Determine the  percent moisture by
      drying overnight at 105 °C.   Allow to cool in a desiccator before
      weighing.  Concentrations of individual analytes will be reported
      relative to the dry weight of soil/sediment.
      EQ. 4
               ^moisture = g of ^t sample - g of dry sample x 10Q
                                   g of wet sample
                                   D-36/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 11
                                                Data Analysis and Calculations
                                                    Qualitative Identification
11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Qualitative Identification

ll.l.i   Identification of Target Compounds

11.1.1.1    The compounds listed in the Target Compound List (TCL) in Exhibit
            C  (Volatiles) shall be identified by an analyst competent in the
            interpretation of mass spectra (see Exhibit A, Section 4.3.1) by
            comparison of the sample mass spectrum to the mass spectrum of a
            standard of the suspected compound.  Two criteria must be
            satisfied to verify the identifications: (1) elution of the sample
            component at the same GC relative retention time as the standard
            component, and (2) correspondence of the sample component and
            standard component mass spectra.

11.1.1.2    For establishing correspondence of the GC relative retention time
            (RRT), the sample component RRT must compare within ±0.06RRT
            units of the RRT of the standard component.  For reference, the
            standard must be run in the same 12-hour time period as the
            sample.   If samples are analyzed during the same 12-hour time
            period as the initial calibration standards, use the RRT values
            from the 50 /xg/L standard.  If co-elution of interfering
            components prohibits accurate assignment of the sample component
            RRT from the total ion chromatogram,  the RRT should be assigned by
            usir  extracted ion current profiles for ions unique to the
            component of interest.

11.1.1.3    For comparison of standard and sample component mass spectra, mass
            spectra obtained on the Contractor's GC/MS are required.   Once
            obtained,  these standard spectra may be used for identification
            purposes,  only if the Contractor's GC/MS meets the daily
            instrument performance requirements for BFB.  These standard
            spectra may be obtained from the run used to obtain reference
            RRTs.

11.1.1.4    The requirements for qualitative verification by comparison of
            mass spectra are as follows:

            •   All  ions present in the standard mass spectra at a relative
                intensity greater than 10.0 percent (most abundant ion in the
                spectrum equals 100.0 percent) must be present in the sample
                spectrum.

            •   The  relative intensities of ions specified above must agree
                within ± 20.0 percent between the standard and sample spectra.
                (Example:  For an ion with an abundance of 50.0 percent in the
                standard spectra,  the corresponding sample abundance must be
                between 30.0 and 70.0 percent).

            •   Ions greater than 10.0 percent in the sample spectrum but not
                present in the standard spectrum must be considered and
                                   D-37/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 11
Data Analysis and Calculations
Qualitative Identification
                accounted for by the analyst making the comparison.  In
                Exhibit A, Task II, the verification process should favor
                false positives.  All compounds meeting the identification
                criteria must be reported with their spectra.  For all
                compounds below the CRQL, report the actual value followed by
                a "J", e.g., "3J".

11.1.1.5    If a compound cannot be verified by all of the criteria in
            11.1.1.4, but in the technical judgment of the mass spectral
            interpretation specialist, the identification is correct, then the
            Contractor shall report that identification and proceed with
            quantitation in Section 11.2.

11.1.2   Identification of Non-Target Compounds

11.1.2.1    A library search shall be executed for non-target sample
            components for the purpose of tentative identification.  For this
            purpose, the NIST/EPA/NIH (May 1992 release or later) and/or Wiley
            (1991 release or later), or equivalent mass spectral library,
            shall be used.

11.1.2.2    Up to 30 organic compounds of greatest apparent concentration not
            listed in Exhibit C for the volatile and semivolatile organic
            fraction, excluding the system monitoring compounds and internal
            standard compounds, shall be tentatively identified via a forward
            search of the NIST/EPA/NIH (May 1992 release or later) and/or
            Wiley (1991 release or later),  or equivalent mass spectral
            library.  The following are not to be reported: 1)  Substances with
            responses less than 10 percent of the internal standard (as
            determined by inspection of the peak areas or height), 2)
            Substances which elute earlier than 30 seconds before the first
            purgeable compound listed in Exhibit C (Volatiles)  or three
            minutes after the last purgeable compound listed in Exhibit C
            {Volatiles) has eluted are not required to be searched in this
            fashion, 3) Carbon dioxide, and 4) Semivolatile TCL compounds
            listed in Exhibit C).   Only after visual comparison of sample
            spectra with the nearest library searches will the mass spectral
            interpretation specialist assign a tentative identification.

11.1.2.3    NOTE:  Computer generated library search routines must not use
            normalizations which would misrepresent the library or unknown
            spectra when compared to each other.

11.1.2.4    Guidelines for making tentative identification:

            •   Relative intensities of major ions in the reference spectrum
                (ions greater than 10.0 percent of the most abundant ion)
                should be present in the sample spectrum.

            •   The relative intensities of the major ions should agree within
                ±20.0 percent.  (Example: For an ion with an abundance of
                50.0 percent of the standard spectra, the corresponding sample
                ion abundance must be between 30.0 and 70.0 percent.)

                                   D-38/VOA                            OLM03.1

-------
                                             Exhibit D Volatiles — Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
            •   Molecular ions present in reference spectrum should be present
                in sample spectrum.

            •   Ions present in the sample spectrum but not in the reference
                spectrum should be reviewed for possible background
                contamination or presence of co-eluting compounds.

            •   Ions present in the reference spectrum but not in the sample
                spectrum should be reviewed for possible subtraction from the
                sample spectrum because of background contamination or co-
                eluting compounds.  Data system library reduction programs can
                sometimes create these discrepancies.

11.1.2.5    If, in the technical judgment of the mass spectral interpretation
            specialist, no valid tentative identification can be made, the
            compound should be reported as unknown.  The mass spectral
            specialist should give additional classification of the unknown
            compound, if possible (i.e.,  unknown aromatic, unknown
            hydrocarbon, unknown acid type,  unknown chlorinated compound) .  If
            probable molecular weights can be distinguished, include them.

11.2  Calculations

11.2.1   Target Compounds

11.2.1.1    Target compounds identified shall be quantified by the internal
            standard method using the equations below.   The internal standard
            used shall be that which is assigned in Table 3.  The relative
            response factor (RRF) from the continuing calibration standard is
            used to calculate the concentration in the  sample.

11.2.1.2    Water
                            „     <-  „•     /r
                            Concentration pg/L  =
                                                 (Ais) (RRF) (V0)

            Where,

            Ax   =  Area of the characteristic ion (EICP)  for  the compound to be
                  measured (see Table 2)
            AJS =  Area of the characteristic ion (EICP)  for  the specific
                  internal standard (see Tables 3 and 4)
            Is   =  Amount of internal standard added in nanograms (ng)
            RRP =  Relative response factor from the ambient  temperature purge
                  of the calibration standard.
            V0   =  Volume of water purged in milliliters  (mL)
            Df   =  Dilution factor.   The dilution factor  for  analysis of water
                  samples for volatiles by this method is defined as the ratio
                  of the number of  milliliters  (mL) of water  purged (i.e., V0
                  above) to the number of mL of the original water sample used
                  for purging.   For example, if 2.0 mL of sample is diluted to


                                   D-39/VOA                            OLM03.0

-------
Exhibit D Volatiles  —  Section 11
Data Analysis and  Calculations
Calculations
                   5 mL  with reagent water and purged, Df = 5 mL/2.0 mL  =  2.5
                   If  no dilution is performed, Df = 1.
11.2.1.3    Low Soil/Sediment
            EQ. 6
           Concentration ug/Kg  (dry weight basis)  =  -      x   s
                                                    (Aia) (RRF) (Ws) (D)

            Where,

            Ax, Is,  Ajs are as given  for  water.
            RRF = Relative response  factor from the heated purge of the
                  calibration standard.
            D   = 100 - % moisture
                        100
            Ws  = Weight of  sample added to the purge tube, in grams  (g)

11.2.1.4    Medium Soil/Sediment

            EQ. 7

                         /    ,      •  ,_,..<    (AJ  (I,) (Vt) (1000) (Df)
         Concentration pg/Kg (Dry weight basis) =   *   s   c
                                                            ^ ^ (£))


            Where,

            Ax,  Is,  AJS are as given for water.
            RRF = Relative response factor  from the ambient temperature purge
                  of the calibration  standard.
            Vt =  Total volume of the methanol  extract in milliliters  (mL).
                  NOTE: This volume is typically 10 mL, even though  only  1 mL
                  is transferred to the vial  in Section 10.1.5.6.
            Va =  Volume of the aliquot of  the  sample methanol extract  (i.e.,
                  sample extract not  including  the methanol added to equal 100
                  /iL) in microliters  (fiL) added to reagent water  for purging.
            Ws =  Weight of soil/sediment extracted, in grams  (g).
            D  =   100 - % moisture
                        100
            Df =  Dilution factor.  The dilution factor for analysis of
                  soil/sediment samples for volatiles by the medium  level
                  method is defined as:

                        u\. most cone, extract used to make dilution + nl clean  solvent
                              nL most cone, extract used to make dilution

                  The dilution factor is equal  to 1.0 in all cases other  than
                  those requiring dilution  of the sample methanol extract
                  (Vt).  Dilution of  the extract is required when the X factor
                  (Section .10.1.5.7)  is  S12.5.  The factor of 1,000 in the
                  numerator converts  the value  of Vt from mL to jtL.


                                   D-40/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
11.2.1.5    For water, low level and medium level soil/sediment samples,
            xylenes (o-,m- and p-isomers)  are to be reported as xylenes
            (total).  Because the o- and p-xylene isomers co-elute on packed
            columns, and the m- and p-xylene isomers co-elute on capillary
            columns, special attention must be given to the quantitation of
            the xylenes.  The relative response factor (RRF) determined in
            Section 9.4.4 is based on the peak that represents the single
            isomer on the GC column used (m-xylene on packed columns, o-xylene
            on capillary columns).  In quantitating sample  concentrations,
            use the areas on both peaks and the RRF from Section 9.4.4.  The
            areas of the two peaks may be summed, and the concentration
            determined, or the concentration represented by each of the two
            peaks may be determined separately, and then summed.  It is
            required that all three xylene isomers be present in the initial
            and continuing calibration standards.

11.2.1.6    The cis and trans stereo isomers of 1,2-dichloroethene are to be
            reported as 1,2-dichloroethene (total).  If the two isomers co-
            elute on the GC column used for analysis, use the area of the co-
            eluting peaks and the RRF determined in Section 9.4.4 to calculate
            the concentration.  If the isomers do not co-elute, use the single
            RRF values determined in Section 9.3.4 to calculate the
            concentration.
11.2.1.7    Secondary ion quantitation is allowed only when there are sample
            matrix interferences with the primary ion.  If secondary ion
            quantitation is performed, document the reasons in the SDG
            Narrative.  A secondary ion cannot be used unless a relative
            response factor is calculated using the secondary ion.

11.2.1.8    The requirements listed in 11.2.1.9 and 11.2.1.10 apply to all
            standards, samples including MS/MSDs, and blanks.

11.2.1.9    It is expected that situations will arise where the automated
            quantitation procedures in the GC/MS software provide
            inappropriate quantitations.  This normally occurs when there is
            compound co-elution, baseline noise, or matrix interferences.  In
            these circumstances the Contractor must perform a manual
            quantitation.  Manual quantitations are performed by integrating
            the area of the quantitation ion of the compound.  This
            integration shall only include the area attributable to the
            specific TCL compound.  The area integrated shall not include
            baseline background noise.  The area integrated shall not extend
            past the point where the sides of the peak intersect with the
            baseline noise.  Manual integration is not to be used solely to
            meet QC criteria, nor is it to be used as a substitute for
            corrective action on the chromatographic system.  Any instances of
            manual integration must be documented in the SDG Narrative.

11.2.1.10   In all instances where the data system report has been edited, or
            where manual integration or quantitation has been performed, the
                                   D-41/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 11
Data Analysis and Calculations
Calculations
            GC/MS operator must identify such edits or manual procedures by
            initialing and dating the changes made to the report, and shall
            include the integration scan range.  In addition, a hardcopy
            printout of the EICP of the quantitation ion displaying the manual
            integration shall be included in the raw data.  This applies to
            all compounds listed in Exhibit C (Volatiles), internal standards
            and system monitoring compounds.

11.2.2   Non-Target Compounds

11.2.2.1    An estimated concentration for non-target compounds tentatively
            identified shall be determined by the internal standard method.
            For quantitation, the nearest internal standard free of
            interferences shall be used.

11.2.2.2    The formulas for calculating concentrations are the same as in
            Sections 11.2.1.2, 11.2.1.3, and 11.2.1.4.   Total area counts  (or
            peak heights) from the total ion chromatograms are to be used  for
            both the compound to be measured and the internal standard.  A
            relative response factor (RRF)  of one (1) is to be assumed.  The
            resulting concentration shall be qualified as "J" (estimated,  due
            to lack of a compound-specific response factor),  and "N"
            (presumptive evidence of presence),  indicating the quantitative
            and qualitative uncertainties associated with this non-target
            component.  An estimated concentration must be calculated for  all
            ten itively identified compounds as /ell as those identified as
            unknowns.

11.2.3   CRQL Calculations

         NOTE: If the adjusted CRQL is less than the CRQL listed in Exhibit C
         (Volatiles), report the CRQL listed in Exhibit C (Volatiles).

11.2.3.1    Water

            EQ. 8

                         Adjusted _ Contract   vx   nf
                           CRQL   ~   CRQL     ~V~0


            Where,

            V0 and  Df are as  given in Equation 5
            Vx = Contract Sample Volume  (5  mL)
                                   D-42/VOA                            OLM03.0

-------
                                             Exhibit  D Volatiles  —  Section 11
                                                Data  Analysis  and Calculations
                                                                   Calculations
11.2.3.2    Low Level Soil/Sediment

            EQ. 9
                         Adjusted _ Contract
                            CRQL       CRQL   A  (w ) (D)
            Where,

            Ws and D are as given in Equation 6
            Wx = Contract Sample Weight (5 g)
11.2.3.3    Medium Level Soil/Sediment

            EQ. 10

                    Adjusted  _ Contract    (l^x) (Vt) (Vy)
                      CRQL        CRQL   x  (ffs) (Vc) (Va) (D)


            Where,

            Vt, Df,  Ws/  Va and  D are  as  given  in  Equation 7
            Wx = Contract Sample Weight (4 g)
            Vv = Contract Soil Aliquot Volume from soil methanol extract  (100
            Vc = Contract Soil Methanol Extract Volume (10,000 fj.'L)

11.2.4    System Monitoring Compound Recoveries

11.2.4.1    Calculate the recovery of each system monitoring  compound  in all
            samples, blanks, matrix spikes and matrix  spike duplictes.
            Determine if the recovery is within limits  (see Table 7),  and
            report on appropriate form.

11.2.4.2    Calculate the concentrations of the system monitoring compounds
            using the same equations as used for target  compounds.

11.2.4.3    Calculate the recovery of each system monitoring  compound  as
            follows:
            EQ. 11
                ^Recovery =  Concentration (amount)  found x 10Q
                            Concentration (amount) spiked
11.2.5   Internal Standard Responses and Retention Times

         Internal standard responses and retention times  in  all  samples  must
         be evaluated during or immediately after data  acquisition.   Compare
         the sample internal standard responses and  retention  times  to the

                                   D-43/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 11
Data Analysis and Calculations
Technical Acceptance Criteria for Sample Analysis

         continuing calibration internal standard response and retention
         times.  For samples analyzed during the same 12-hour time period as
         the initial calibration standards,  compare the internal standard
         responses and retention times against the 50 /xg/L calibration
         standard.  The extracted ion current profile (EICP)  of the internal
         standards must be monitored and evaluated for each sample, blank,
         matrix spike and matrix spike duplicate.

11.3  Technical Acceptance Criteria for Sample Analysis

11.3.1   The samples must be analyzed on a GC/MS system meeting the BFB,
         initial calibration, continuing calibration, and blank technical
         acceptance criteria.

11.3.2   The sample must be analyzed within the contract holding time.

11.3.3   The sample must have an associated method blank meeting the blank
         technical acceptance criteria.

11.3.4   The percent recovery of each of the system monitoring compounds in
         the sample must be within the acceptance windows in  Table 7.

11.3.5   The EICP area for each of the internal standards must be within the
         inclusive range of -50.0 percent and +100.0 percent  of the response
         of the internal standards in the most recent continuing calibration
         analysis.

11.3.6   The retention time shift for each of the internal standards must be
         within ±0.50 minutes (30 seconds)  between the sample and the most
         recent continuing calibration standard analysis.

11.3.7   The relative retention time (RRT) of the system monitoring compound
         in a sample must be within ±0.06 (RRT) units of its  relative
         retention time in the continuing calibration standard.

11.3.8   Excluding those ions in the solvent front, no ion may saturate the
         detector.  No target compound concentration may exceed the upper
         limit of the initial calibration range unless a more diluted aliquot
         of the sample is also analyzed according to the procedures in Section
         10.1.6.

11.3.9   The Contractor must demonstrate that there is no carryover from a
         contaminated sample before data from subsequent analyses may be
         submitted.  After a sample that contains a target compound at a level
         exceeding the initial calibration range, the Contractor must either:

         •   Analyze an instrument blank immediately after the contaminated
             sample.  If an autosampler is used, an instrument blank must also
             be analyzed using the same purge inlet that was  used for the
             contaminated sample.  The instrument blanks must meet the
             technical acceptance criteria for blank analysis  (see Section
             12.1.4), or
                                   D-44/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 11
                                                Data Analysis and Calculations
                                         Corrective Action for Sample Analysis
         •   Monitor the sample analyzed immediately after the contaminated
             sample for all compounds that were in the contaminated sample and
             that exceeded the calibration range.  The maximum contamination
             criteria are as follows:  the sample must not contain a
             concentration above the CRQL for the target compounds that
             exceeded the limits in the contaminated sample.  If an auto
             sampler is used, the next sample analyzed using the same purge
             inlet that was used for the contaminated sample also must meet
             the maximum contamination criteria. If the maximum criteria were
             exceeded, then all samples affected by the carryover must be
             reanalyzed at no additional cost to the Agency.

11.4  Corrective Action for Sample Analysis

11.4.1   Sample technical acceptance criteria must be met before data are
         reported.  Samples contaminated from laboratory sources or sample
         results which failed to meet the sample technical acceptance criteria
         require reanalysis at no additional cost to the Agency.

11.4.2   Corrective actions for failure to meet instrument performance checks,
         initial and continuing calibration, and method blanks must be
         completed before the analysis of samples.

11.4.3   Corrective action for system monitoring compounds and internal
         standard compounds that fail to meet acceptance criteria.

11.4.3.1    If any of the system monitoring compounds and internal standard
            compounds fail to meet acceptance criteria:

            •   Check all calculations,  instrument logs,  the system monitoring
                compound and internal standard compound spiking solutions,  and
                the instrument operation.   If the calculations were incorrect,
                correct the calculations and verify that the system monitoring
                compound recoveries and internal standard compound responses
                meet acceptance criteria.

            •   If the instrument logs indicate that the incorrect amount of
                system monitoring compound or internal standard compound
                spiking solution was added,  then reanalyze the sample after
                adding the correct amount of system monitoring compound and
                internal standard spiking solutions.

            •   If the system monitoring compound spiking solution or internal
                standard compound spiking solution was improperly prepared,
                concentrated,  or degraded,  re-prepare the solutions and
                reanalyze the samples.

            •   If the instrument malfunctioned, correct the instrument
                problem and reanalyze the sample.   If the instrument
                malfunction affected the calibration, recalibrate the
                instrument before reanalyzing the sample.   Verify that the
                system monitoring compound recoveries meet acceptance
                criteria.

                                   D-45/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 11
Data Analysis and Calculations
Corrective Action for Sample Analysis

11.4.3.2    If the above actions do not correct the problem,  then the problem
            may be due to a sample matrix effect.   To determine if there was a
            matrix effect, take the following corrective action steps:

            •   Reanalyze the sample.  EXCEPTION:  If system monitoring
                compound recoveries or internal standard compound responses in
                a sample used for a matrix spike or matrix spike duplicate
                were outside the acceptance criteria, then it should be
                reanalyzed only if system monitoring compound recoveries and
                internal standard compound responses met acceptance criteria
                in both the matrix spike and matrix spike duplicate analyses.

            •   If the system monitoring compound recoveries and the internal
                standard compound responses meet the acceptance criteria in
                the reanalyzed sample, then the problem was within the
                Contractor's control.  Therefore,  submit data only from the
                reanalysis.

            •   If the system monitoring compound recoveries and/or the
                internal standard compound responses fail to meet the
                acceptance windows in the reanalysis, then submit data from
                both analyses.  Distinguish between the initial analysis and
                the reanalysis on all deliverables using the suffixes on
                Exhibit B.

11.4.4   Corrective action for system monitoring compounds relative retention
         times/internal standard compounds retention times outside acceptance
         criteria.

11.4.4.1    If the system monitoring compounds relative retention times or
            internal standard compounds retention times are not within their
            acceptance criteria, check the instrument for malfunctions.  If
            the instrument malfunctioned,  correct the instrument problem and
            reanalyze the sample.  If the instrument malfunction affected the
            calibration, recalibrate the instrument before reanalyzing the
            samples.

11.4.4.2    If the above actions do not correct the problem,  then the problem
            may be due to a sample matrix effect.   To determine if there was a
            matrix effect, take the following corrective action steps:

            •   Reanalyze the sample.  EXCEPTION:  If the system monitoring
                compounds relative retention times or internal standard
                compounds retention times in a sample used for a matrix spike
                or matrix spike duplicate were outside the acceptance
                criteria, then it should be reanalyzed only if the system
                monitoring compounds and internal standard compounds retention
                times were within the acceptance criteria in both the matrix
                spike and matrix spike duplicate analyses.

            •   If the system monitoring compounds relative retention times
                and internal standard compounds retention times are within the
                acceptance criteria, then the problem was within the
                Contractor's control.  Therefore,  submit only data from the

                                   D-46/VOA                            OLM03.1

-------
                             Exhibit D Volatiles — Section 11
                                Data Analysis and Calculations
                         Corrective Action for Sample Analysis


reanalysis when the system monitoring compounds relative
retention times and the internal standard compounds retention
times are within the acceptance limits.

If the system monitoring jompounds relative retention times or
the internal standard compounds retention times are outside
the acceptance criteria in the reanalysis, then submit data
from both analyses.  Distinguish between the initial analysis
and the reanalysis on all deliverables, using the suffixes in
Exhibit B.
                   D-47/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 12
Quality Control
Blank Analyses

12.0  QUALITY CONTROL

12.1  Blank Analyses

12.1.1   Summary — There are three different types of blanks required by this
         method.

12.1.1.1    METHOD BLANK - a volume of a clean reference matrix (reagent water
            for water samples or a purified solid matrix for soil/sediment
            samples) that is carried through the entire analytical procedure.
            The volume or weight of the reference matrix must be approximately
            equal to the volume or weight of samples associated with the
            blank.  The purpose of a method blank is to determine the levels
            of contamination associated with the processing and analysis of
            samples.

12.1.1.2    STORAGE BLANK - upon receipt of the first samples in an SDG, two
            40.0 mL screw-cap volatile vials with a PTFE-faced silicone septum
            are filled with reagent water (80 mL total).  The vials are stored
            with the samples in the SDG under the same conditions.  After all
            samples in the SDG have been analyzed,  the storage blank is
            analyzed.   The storage blank indicates whether contamination may
            have occurred during storage of samples.

12.1.1.3    INSTRUMENT BLANK - a 5.0 mL aliquot of reagent water that is
            carried through the entire analytical procedure.  Instrument
            blanks are analyzed after a sample/dilution which contains a
            target compound exceeding the initial calibration range.  The
            results from the instrument blank analysis indicate whether there
            is contamination from a previous sample.

12.1.2   Frequency of  Blank Analyses

12.1.2.1    The method blank must be analyzed at least once during every 12-
            hour time  period on eacn GC/MS system used for volatile analysis
            (see Section 9.2.2 for the definition of the 12-hour time period).

12.1.2.2    The method blank must be analyzed after the continuing calibration
            and before any samples,  including matrix spike/matrix spike
            duplicates,  dilutions or storage blanks are analyzed.   The method
            blank must be analyzed after the initial calibration sequence if
            samples are analyzed before the 12-hour period expires.  A method
            blank must be analyzed in each 12-hour time period in which
            samples,  including dilutions,  matrix spikes/matrix spike
            duplicates,  and storage blanks are analyzed.

12.1.2.3    A minimum  of one storage blank must be analyzed per SDG after all
            samples for that SDG have been analyzed.

12.1.2.4    The Contractor must demonstrate that there is no carryover from
            contaminated samples before data from subsequent analyses may be
            used.   Samples/dilutions may contain target compounds at levels
            exceeding  the initial calibration range.  An instrument blank must
                                   D-48/VOA                            OLM03.0

-------
                                             Exhibit  D Volatiles  —  Section  12
                                                                Quality  Control
                                                                Blank Analyses

             be  analyzed  after the  sample that exceeds the  calibration range
             (also  in  the same purge  inlet  if an autosampler  is  used)  or a
             sample that  meets the  maximum  contamination  criteria  in  Section
             11.3.8 must  be analyzed.  For  these purposes,  if the  instrument
             blank  meets  the technical acceptance criteria  for blank  analyses
             or  the sample meets the  max_mum contamination  criteria,  the system
             is  considered to be uncontaminated.  If the  instrument blank or
             sample does  not meet the criteria (i.e.,  contaminated),  the system
             must be decontaminated.  Until an instrument blank  meets  the blank
             technical acceptance criteria  or a sample meets  the maximum
             contamination criteria in Section 11.3.8, any  samples analyzed
             since  the original contaminated sample will  require reanalysis at
             no  additional cost to  the Agency.  NOTE:  Only  the instrument blank
             which  demonstrates that  there was no carryover from the previous
             sample or the instrument blank that demonstrates that the system
             is  clean  (Section 12.1.4) needs to be reported.  Instrument blanks
             analyzed during the instrument decontamination process which
             exceed the requirements  listed in Section 12.1.4 do not  need to  be
             reported.

12.1.3   Procedure for Blank Analyses

12.1.3.1     For water samples, a volatile method blank consists of a  5  mL
             volume of reagent water  (Section 7.1.1) spiked with 10 /xL of the
             system monitoring compound spiking solution  (Section 7.2.4.1) and
             10 jiL  of the  internal  standard spiking solution  (Section  7.2.4.3)
             and carried through the  analytical procedure.

12.1.3.2     For low level soil/sediment samples, a volatile  method blank
             consists of  5 g of a purified solid matrix added to 5 mL  of
             reagent water that has been spiked with 10 iiL  each  of the system
             monitoring compound spiking solution and  the internal standard
             spiking solution.  The method blank is then carried through the
             analytical procedure.

12.1.3.3     For medium level soil/sediment samples, a volatile  method blank
             consists of 4 g of a purified solid matrix added to 10 mL of
             methanol and  extracted for two minutes.   A 100 /xL aliquot of  the
            methanol is added to reagent water and spiked  with  10 /xL of  the
             internal standard spiking solution and 10 /xL of  the system
            monitoring compound spiking solution and  taken through the
             analytical procedure.

12.1.3.4     Storage/instrument blanks consist of a 5 mL volume of reagent
            water  (Section 7.1.1)   spiked with 10 /uL of the system monitoring
             compound spiking solution (Section 7.2.4.1)  and  10  iiL of the
             internal standard spiking solution (Section 7.2.4.3) and carried
            through the analytical procedure.

12.1.3.5     Identify and quantitate analytes according to  Section 11.0.
                                   D-49/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 12
Quality Control
Blank Analyses

12.1.4   Technical Acceptance Criteria for Blank Analyses

12.1.4.1    All blanks must be analyzed on a GC/MS system meeting the BFB,
            initial calibration and continuing calibration technical
            acceptance criteria and at the frequency described in Section
            12.1.2.

12.1.4.2    The percent recovery of each of the system monitoring compounds in
            a blank must be within the acceptance windows in Table 7.

12.1.4.3    The EICP area for each of the internal standards in a blank must
            be within the inclusive range of -50.0 percent and +100.0 percent
            of the response of the internal standards in the most recent
            continuing calibration analysis.

12.1.4.4    The retention time shift for each of the internal standards in a
            blank must be within ±0.50 minutes (30 seconds)  of its retention
            time in the most recent continuing calibration standard analysis.

12.1.4.5    The concentration of each target compound found in the blank must
            be less than its CRQL listed in Exhibit C (Volatiles), except for
            methylene chloride which must be less than 2.5 times its CRQL,  and
            acetone and 2-butanone, which must be less than 5 times the CRQL.

12.1.5   Corrective Action for Blank Analyses

12.1.5.1    It is the Contractor's responsibility to ensure that method
            interferences caused by the contaminants in solvent, reagents,
            glassware, laboratory air and other sample storage and processing
            hardware that lead to discrete artifacts and/or elevated baselines
            in gas chromatograms be eliminated.  If a Contractor's blanks
            exceed the criteria in Section 12.1.4.5, the Contractor must
            consider the analytical system to be out of control.  The source
            of the contamination must be investigated and appropriate
            corrective measures must be taken and documented before further
            analysis proceeds.

12.1.5.2    Any method blank or instrument blank that fails to meet the
            technical acceptance criteria for blank analyses must be
            reanalyzed at no additional cost to the Agency.   Furthermore, all
            samples, including MS/MSD samples, processed within the 12-hour
            period with a method blank or instrument blank that does not meet
            the technical acceptance criteria for blanks will require
            reanalysis at no additional cost to the Agency.

12.1.5.3    If the storage blank does not meet the technical acceptance
            criteria for blank analyses in Sections 12.1.4.1 through 12.1.4.4,
            correct system problems and reanalyze the storage blank.  If the
            storage blank does not meet the criteria in Section 12.1.4.5,
            reanalyze the storage blank to determine whether the contamination
            occurred during storage or during the analysis.   If, upon
            reanalysis, the storage blank meets the criteria in Section
            12.1.4.5, the problem occurred during the analysis and the
            reanalyzed storage blank results must be reported.  If upon

                                   D-50/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 12
                                                               Quality Control
                                  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

            reanalysis, the storage blank did not meet the criteria in Section
            12.1.4.5, the problem occurred during storage.  The laboratory
            manager or his/her designee must address the problem in the SDG
            Narrative and discuss the corrective actions implemented to
            prevent future occurrences.  NOTE: A copy of the storage blank
            data must also be retained by the Contractor and be made available
            for inspection during on-site laboratory evaluations.

12.2  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.1   Summary of MS/MSD

         In order to evaluate the effects of the sample matrix on the methods
         used for volatile analyses, the Agency has prescribed a mixture of
         volatile target compounds to be spiked into two aliquots of a sample,
         and analyzed in accordance with the appropriate method.

12.2.2   Frequency of MS/MSD

12.2.2.1    A matrix spike and matrix spike duplicate must be performed for
            each group of samples of a similar matrix for the following,
            whichever is most frequent:

            •   Each SDG (not to exceed 20 field samples), or

            •   Each matrix within an SDG, or

            •   Each group of samples of a similar concentration level (soils
                only).

12.2.2.2    As a part of the Agency's QA/QC program, water rinsate samples
            and/or field/trip blanks (field QC) may accompany soil/sediment
            samples and/or water samples that are delivered to a laboratory
            for analysis.  The Contractor shall not perform MS/MSD analysis on
            any of the field QC samples.

12.2.2.3    If the EPA Region designates a sample to be used as an MS/MSD,
            then that sample must be used.  If there is insufficient sample,
            less than the required amount, remaining to perform an MS/MSD,
            then the Contractor shall choose another sample to perform an
            MS/MSD analysis.  At the time the selection is made, the
            Contractor shall notify the Region (through SMO) that insufficient
            sample was received and identify the EPA sample selected for the
            MS/MSD analysis.  The rationale for the choice of a sample other
            than the one designated by the Region shall be documented in the
            SDG Narrative.

12.2.2.4    If there is insufficient sample remaining in any of the samples in
            an SDG to perform an MS/MSD, then the Contractor shall immediately
            contact SMO to inform them of the problem.  SMO will contact the
            Region for instructions. The Region will either approve that no
            MS/MSD is required, or require that a reduced sample aliquot be
            used for the MS/MSD analysis.  SMO will notify the Contractor of
                                   D-51/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate (MS/MSD)

            the resolution.  The Contractor shall document the decision in the
            SDG Narrative.

12.2.2.5    The Contractor will not be paid for MS/MSD analysis performed at a
            greater frequency than required by the contract.   If it appears
            that the Region has requested MS/MSD analysis at  a greater
            frequency than required by the contract,  the Contractor shall
            contact SMO.  SMO will contact the Region to determine which
            samples should have an MS/MSD performed on them.   SMO will notify
            the Contractor of the Region's decision.   The Contractor shall
            document the decision in the SDG Narrative.

12.2.2.6    When a Contractor receives only a Performance Evaluation (PE)
            sample(s), no MS/MSD shall be performed within that SDG.

12.2.2.7    When a Contractor receives a PE sample as part of a larger SDG, a
            sample other than the PE sample must be chosen for the MS/MSD when
            the Region did not designate samples to be used for this purpose.
            If the PE sample is received as an ampulated standard extract, the
            ampulated PE sample is not considered to be another matrix type.

12.2.3   Procedure for Preparing MS/MSD

12.2.3.1    Water

            To prepare a matrix spike and matrix spike duplicate for water
            samples, add 10 /iL of the matrix spike solution (Section 7.2.4.2)
            to each of the 5 mL aliquots of the sample chosen for spiking.
            Process samples according to Sections 10.1.3.7 through 10.1.3.12.
            Disregarding any dilutions, this is equivalent to a concentration
            of 50 ^g/L of each matrix spike compound.

12.2.3.2    Soil - Low Level

            To prepare a matrix spike and matrix spike duplicate for low level
            soil/sediment samples, add 10 fiL of the matrix spike solution to
            the 5 mL of spiked reagent water added to each of the two aliquots
            of the soil/sediment from the sample chosen for spiking.  Process
            samples according to Sections 10.1.4.7 through 10.1.4.12.  The
            concentration for a 5 g sample should be equivalent to 50 /*g/kg of
            each matrix spike compound.

12.2.3.3    Soil - Medium Level

12.2.3.3.1     To prepare a matrix spike and matrix spike duplicate for medium
               level soil/sediment samples, add 9 mL of methanol and 1 mL of
               matrix spike solution to each of the two aliquots of the
               soil/sediment sample chosen for spiking.   Process samples
               according to Sections 10.1.5.6 through 10.1.5.10.  This results
               in a 6,200 /ig/kg concentration of each matrix  spike compound
               when added to a 4 g sample.  Add a 100 pL aliquot of this
               extract to 5 mL of water for purging (as per Sections 10.1.5.8
               through 10.1.5.9).
                                   D-52/VOA                            OLM03.0

-------
                                             Exhibit D Volatiles — Section 12
                                                               Quality Control
                                  Matrix Spike/Matrix Spike Duplicate (MS/MSD)
12.2.3.3.2     NOTE:  Before performing an MS/MSD analysis, analyze the sample
               used for MS/MSD.  If the sample analysis required dilution, the
               aliquots for the MS/MSD can be prepared at the same dilution as
               the least diluted analysis for which the sample results will be
               reported to the Agency.  Sample dilutions must be performed in
               accordance with Section 10.1.6.  Do not further dilute MS/MSD
               samples to get either spiked or non-spiked analytes within
               calibration range.

12.2.4   Calculations for MS/MSD

12.2.4.1    Calculate the concentrations of the matrix spike compounds using
            the same equations as used for target compounds (Equations 5, 6
            and 7).  Calculate the recovery of each matrix spike compound as
            follows:

            EQ. 12
                     Matrix Spike Recovery = -	— x  100
            Where,

            SSR = Spiked sample result
            SR = Sample result
            S? = Spike added
12.2.4.2    Calculate the relative percent difference (RPD) of the recoveries
            of each compound in the matrix spike and matrix spike duplicate as
            follows:

            EQ. 13
                           RPD =       -        x 100
                                 — (MSR + MSDR)
            Where,

            MSR = Matrix spike recovery
            MSDR = Matrix spike duplicate recovery

            The vertical bars in the formula above indicate the absolute value
            of the difference, hence RPD is always expressed as a positive
            value.
                                   D-53/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 12
Quality Control
Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.5   Technical Acceptance Criteria for MS/MSD

12.2.5.1    All MS/MSDs must be analyzed on a GC/MS system meeting the BFB,
            initial calibration and continuing calibration technical
            acceptance criteria,  the blank technical acceptance criteria, and
            at the frequency described in Section 12.2.2.

12.2.5.2    The MS/MSD must be analyzed within the contract holding time.

12.2.5.3    The retention time shift for each of the internal standards in the
            MS/MSD must be within ±0.50 minutes (30 seconds) of its retention
            time and the most recent continuing calibration standard analysis.

12.2.5.4    The limits for matrix spike compound recovery and RPD are given in
            Table 8.  As these limits are only advisory,  no further action by
            the laboratory is required.  However,  frequent failures to meet
            the limits for recovery or RPD warrant investigation by the
            laboratory, and may result in questions from the Agency.

12.2.6   Corrective Action for MS/MSD

         Any MS/MSD that does not meet the technical acceptance criteria for
         MS/MSD must be reanalyzed at no additional cost to the Agency.
                                   D-54/VOA                            OLM03.0

-------
                                       Exhibit D Volatiles — Sections 13 - 16
           Method Performance/Pollution Prevention/Waste Management/References
13.0  METHOD PERFORMANCE

      Not applicable.

14.0  POLLUTION PREVENTION

14.1  Pollution prevention encompasses any technique that reduces or
      eliminates the quantity or toxicity of waste at the point of generation.
      Numerous opportunities for pollution prevention exist in laboratory
      operation.  The EPA has established a preferred hierarchy of
      environmental management techniques that places pollution prevention as
      the management option of first choice.  Whenever feasible, laboratory
      personnel should use pollution prevention techniques to address their
      waste generation.  When wastes cannot be feasibly reduced at the source,
      the Agency recommends recycling as the next best option.

14.2  For information about pollution prevention that may be applicable to
      laboratories and research institutions consult Less is Better:
      Laboratory Chemical Management for Waste Reduction, available from the
      American Chemical Society's Department of Government Relations and
      Science Policy, 1155 16th Street, N.W., Washington B.C., 20036, (202)
      872-4477.

15.0  WASTE MANAGEMENT

      The Environmental Protection Agency requires that laboratory waste
      management practices be conducted consistent with all applicable rules
      and regulations.  The Agency urges laboratories to protect the air,
      water, and land by minimizing and controlling all releases from hoods
      and bench operations, complying with the letter and spirit of any sewer
      discharge permits and regulations, and by complying with all solid and
      hazardous waste regulations,  particularly the hazardous waste
      identification rules and land disposal restrictions.  For further
      information on waste management consult The Waste Management Manual for
      Laboratory Personnel, available from the American Chemical Society at
      the address listed in Section 14.2.

16.0  REFERENCES

      Not applicable.
                                   D-55/VOA                            OLM03.0

-------
Exhibit D Volatiles — Section 17
Tables/Diagrams/Flowcharts
17.0  TABLES/DIAGRAMS/FLOWCHARTS
                                   Table  1

                   BFB Key  Ions and  Ion Abundance Criteria



              Mass     Ion  Abundance Criteria

              50       8.0-40.0  percent  of  mass  95
              75       30.0-66.0 percent of mass 95
              95       base peak, 100  percent  relative  abundance
              96       5.0-9.0 percent of mass 95 (see  note)
              173      less than 2.0 percent of  mass  174
              174      50.0-120.0 percent of mass 95
              175      4.0-9.0 percent of mass 174
              176      93.0-101.0 percent of mass 174
              177      5.0-9.0 percent of mass 176
         NOTE:  All ion abundances must be normalized to m/z 95,  the nominal
         base peak,  even though the ion abundance of m/z 174 may be up to
         120.0 percent that of m/z 95.
                                   D-56/VOA                            OLM03.0

-------
                                             Exhibit  D Volatiles  —  Section 17
                                                   Tables/Diagrams/Flowcharts
                                   Table 2

              Characteristic Ions for Volatile Target Compounds
Analyte
                                      Primary
                                   Quantitation
                                        Ion
Secondary Ion(s)
Chloromethane
Bromomethane
Vinyl chloride
Chloroethane
Methylene chloride
Acetone
Carbon disulfide
1, 1-Dichloroethene
1 , 1-Dichloroethane
1,2-Dichloroethene (total)
Chloroform
1 , 2-Dichloroethane
2-Butanone
1,1, 1-Tr ichloroethane
Carbcn Tetrachloride
Bromodichloromethane
1, 1,2,2-Tetrachloroethane
1,2-Dichloropropane
trans-1 , 3-Dichloropropene
Trichloroethene
Dibromochloromethane
1, 1,2-Trichloroethane
Benzene
cis-1 , 3-Dichloropropene
Bromof orm
2-Hexanone
4-Methyl-2-pentanone
Tetrachloroethene
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Total Xylenes
50
94
62
64
84
43
76
96
63
96
83
62
43*
97
117
83
83
63
75
130
129
97
78
75
173
43
43
164
91
112
106
104
106
52
96
64
66
49,
58
78
61,
65,
61,
85
64,
57
99,
119
85
85,
65,
77
95,
208
83,
—
77
171
58,
58,
129
92
114
91
78,
91




51, 86


98
83, 85, 98, 100
98

100, 98

117, 119
, 121

131, 133, 166
114

97, 132
, 206
85, 99, 132, 134


, 175, 250, 252, 254, 256
57, 100
100
, 131, 166



103

* m/z 43 is used for quantitation of 2-Butanone, but m/z 72 must be present
for positive identification.
                                    D-57/VOA
                                                                       OLM03.0

-------
Exhibit D Volatiles — Section 17
Tables/Diagrams/Flowcharts
                                    Table  3

        Volatile  Internal  Standards  with Corresponding  Target  Compounds
           and  System Monitoring  Compounds Assigned  for Quantitation
Bromochloromethane
   1,4-Difluorobenzene
Chlorobenzene-dj
Chloromethane
Bromomethane
Vinyl Chloride
Chloroethane

Methylene Chloride

Acetone
Carbon Disulfide
1,1-Dichloroethene
1,1-Dichloroethane
1,2-Dichloroethene (tot.)

Chloroform
1,2-Dichloroethane
2-Butanone
l,2-Dichloroethane-d4 (SMC)
   1,1,1-Trichloroethane
   Carbon Tetrachloride
   Bromodichloromethane
   1,2-Dichloropropane

   trans-1,3
     Dichloropropene
   Trichloroethene
   Dibromochloromethane
   1,1,2-Trichloroethane
   Benzene
   cis-1,3-Dichloropropene

   Bromoform
2-Hexanone
4-Methyl-2-Pentanone
Tetrachloroethene
1,1,2,2-
 Tetrachloroethane
Toluene

Chlorobenzene
Ethylbenzene
Styrene
Xylene (total)
4-Bromofluorobenzene
(SMC)
Toluene-d8 (SMC)
(SMC)  = system monitoring compound
                                   Table 4

           Characteristic Ions for System Monitoring Compounds and
      Internal Standards for Volatile Organic Compounds with CAS Numbers
Compound
   Primary
Quantitation Secondary
     Ion       Ion(s)
   CAS Number
4-Bromofluorobenzene
1,2-Dichloroethane-d4
Toluene-ds
 SYSTEM MONITORING COMPOUNDS
     95      174, 176          460-00-4
     65      102               17060-07-0
     98      70, 100           2037-26-5
Bromochloromethane
1,4-Difluorobenzene
Chlorobenzene-dj
      INTERNAL  STANDARDS
    128      49, 130, 51
    114      63, 88
    117      82, 119
   74-97-5
   540-36-3
   3114-55-4
                                   D-58/VOA
                                              OLM03.0

-------
                                    Exhibit D Volatiles — Section 17
                                           Tables/Diagrams/Flowcharts
                           Table  5

Relative Response Factor Criteria for Initial and Continuing
          Calibration  of  Volatile Organic Compounds
Volatile
Compound
Chloromethane
Bromometh ine
Vinyl chloride
Chloroethane
Methylene chloride
Acetone
Carbon disulfide
1, 1-Dichloroethene
1, 1-Dichloroethane
1,2-Dichloroethene (total)
Chloroform
1 , 2-Dichloroethane
2-Butanone
1, 1, 1-Trichloroethane
Carbon tetrachloride
Bromodichloromethane
1 , 2-Dichloropropane
cis-l,3-Dichloropropene
Trichloroethene
Dibromochloromethane
1,1, 2-Trichloroethane
Benzene
trans-1 , 3-Dichloropropene
Bromoform
4-Methy 1-2 -pent anone
2-Hexanone
Tetrachloroethene
1,1,2, 2-Tetrachloroethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Xylenes (total)
SYSTEM MONITORING COMPOUNDS
Bromof luorobenzene
Toluene-dg
1 , 2-Dichloroethane-d4
Minimum
RRF
0.010
0.100
0.100
0.010
0.010
0.010
0.010
0.100
0.200
0.010
0.200
0.100
0.010
0.100
0.100
0.200
0.010
0.200
0.300
0.100
0.100
0.500
0.100
0.100
0.010
0.010
0.200
0.300
0.400
0.500
0.100
0.300
0.300

0.200
0.010
0.010
Maximum
%RSD
none
20.5
20.5
none
none
none
none
20.5
20.5
none
20.5
20.5
none
20.5
20.5
20.5
none
20.5
20.5
20.5
20.5
20.5
20.5
20.5
none
none
20.5
20.5
20.5
20.5
20.5
20.5
20.5

20.5
none
none
Maximum
%Diff
none
±25.0
±25.0
none
none
none
none
±25.0
±25.0
none
±25.0
±25.0
none
±25.0
±25.0
±25.0
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
none
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0

±25.0
none
none
                          D-59/VOA
OLM03.0

-------
Exhibit D Volatiles — Section 17
Tables/Diagrams/Flowcharts
                                   Table  6

                             The "X"  Factor Table
      X Factor
                        Estimated
                        Concentration Range
                                      Take This Volume of
                                      Methanol Extract2
      0.25 - 5.0
      0.5 - 10.0
      2.5 - 50.0
      12.5 - 250
               500 - 10,000
               1000 - 20,000
               5000 - 100,000
               25,000 - 500,000
100
50
10
100 of 1/50 dilution3
         Calculate appropriate dilution factor for concentrations exceeding
         those in the table.
            Actual  concentration  ranges  could be  10 to  20 times  higher  than
         this if the compounds are halogenated  and the  estimates are from
         GC/FID.

         2   The  volume  of  methanol  added to  the 5  mL of water being  purged
         should  be kept constant.  Therefore,  add to  the  5  mL  syringe whatever
         volume  of methanol is necessary to  maintain  a  volume  of 100 /xL added
         to the  syringe.
3  Dilute an aliquot  of the methanol  extract  and then  take 100
analysis.
                                                                           for
                                   D-60/VOA
                                                              OLM03.0

-------
                                       Exhibit D Volatiles — Section  17
                                              Tables/Diagrams/Flowcharts
                              Table 7

            System Monitoring Compound Recovery Limits



   Compound                   %  Recovery  Water  % Recovery Soil

   Toluene-d8                      88-110            84-138
   Bromofluorobenzene              86-115            59-113
   l,2-Dichloroethane-d4           76-114            70-121
                              Table 8

                     Matrix Spike Recovery and
                Relative Percent Difference Limits
Compound             % Recovery    RPD       % Recovery    RPD
                     Water         Water     Soil          Soil

1,1-Dichloroethene   61-145        14        59-172        22
Trichloroethene      71-120        14        62-137        24
Benzene              76-127        11        66-142        21
Toluene              76-125        13        59-139        21
Chlorobenzene        75-130        13        60-133        21
                             D-61/VOA                            OLM03.1

-------
Exhibit D  — Volatiles Appendix A
Screening  of Hexadecane Extracts for Volatiles
APPENDIX A  -  SCREENING OF HEXADECANE EXTRACTS FOR VOLATILES

1.0    SCOPE AND APPLICATION

1.1    The hexadecane extraction and screening methods for purgeables described
       in this section are designed to aid the analyst in deciding whether a
       soil  sample is low or medium level in order to prevent saturation of the
       purge and trap system and/or the GC/MS system.  These or other screening
       methods should be used, particularly if there is some doubt about the
       level of organics in a sample.  This is especially true in soil/sediment
       analysis.  Water samples may also be screened to determine an
       appropriate dilution factor for analysis.

1.2    These extractions and preparation procedures were developed for rapid
       screening of water and soil/sediment samples from hazardous waste sites.
       The design of the methods thus does not stress efficient recoveries or
       low limits of guantitation.  Rather, the procedures were designed to
       screen at moderate recovery and sufficient sensitivity for a broad
       spectrum of organic chemicals.  The results of the analyses thus may
       reflect only a minimum of the amount actually present in some samples.
       This  is especially true if water soluble solvents are present.

2.0    SUMMARY OF METHODS

2.1    Sample Preparation

2.1.1    Water

         A  40 mL aliquot of sample is extracted with 2 mL of hexadecane.  This
         provides a minimum guantitation limit (MQL) as follows:

                   Compounds                  MOL (ua/D
                   non-halogenated aromatics  40-50
                   halogenated methanes       80-1000
                   halogenated ethanes        400-500

2.1.2    Soil/Sediment

         40 mL of reagent water are added to 10 g (wet weight)  of
         soil/sediment and shaken.  The water phase is in turn extracted with
         2 mL of hexadecane.   This provides a minimum quantitation limit of
         approximately four times higher than those listed for water.

2.2   GC/FID Screening

      The hexadecane extracts of water and soil/sediment are screened on a gas
      chromatograph/flame ionization detector (GC/FID).   The results of the
      screen will determine if volatile organics are to be analyzed by low or
      medium level GC/MS procedures if the sample is a soil/sediment, or to
      determine the appropriate dilution factor if the sample is water.  Note:
      The flame ionization detector varies considerably in sensitivity when
      comparing aromatics and halogenated methanes and ethanes. Halomethanes
      are approximately 20x less sensitive than aromatics and haloethanes are


                                   D-62/VOA                            OLM03.0

-------
                                             Exhibit D — Volatiles Appendix A
                                 Screening  of Hexadecane Extracts  for Volatiles
       approximately  lOx  less  sensitive  than  aromatics.  Low molecular weight,
       water  soluble  solvents  (e.g.,  alcohols and ketones, will not be
       extracted  from the water,  and  therefore will not be detected by the
       GC/FID.

 3.0    INTERFERENCES

       Method interferences may be  caused by  contaminants in solvents,
       reagents,  glassware, and other sample  processing hardware that lead to
       discrete artifacts and/or  elevated baselines in the total ion current
       profiles.  All  of  these materials must be routinely demonstrated to be
       free from  interferences under  the conditions of the analysis by running
       laboratory reagent blanks.   Matrix interferences may be caused by
       contaminants that  are coextracted from the sample.  The extent of matrix
       interferences will vary considerably from source to source depending
       upon the nature and diversity  of  the site being sampled.

 4.0    SAFETY

       The toxicity or carcinogenicity of each reagent used in this method has
       not been precisely determined,  however, each chemical should be treated
       as a potential health hazard.   Exposure to these reagents should be
       reduced to the lowest possible  level.   The laboratory is responsible for
       maintaining a current awareness file of OSHA regulations regarding the
       safe handling of the chemicals  specified in this method.  A reference
       file of data handling sheets should be made available to all personnel
       involved in these  analyses.  Use  all reagents in fume hoods whenever
       possible.  Always  wear safety  glasses  or a shield for eye protection,
       protective clothing and observe proper mixing when working with these
       reagents.

 5.0    EQUIPMENT AND SUPPLIES

       Brand names,  suppliers, catalog and part numbers are for illustrative
       purposes only.  No endorsement  is implied.  Equivalent performance may
       be achieved using  equipment  and supplies other than those specified
       here, but demonstration of equivalent performance meeting the
       requirements of this SOW is  the responsibility of the Contractor.

 5.1   Glassware

 5.1.1    Syringes - 0.5 mL

 5.1.2    Vials and Caps  - 2 mL capacity for GC autosampler

 5.1.3    Pasteur Pipets  - disposable

 5.1.4    Centrifuge Tube - 50 mL with ground glass  stopper or Teflon-lined
         screw cap.

5.1.5    Volumetric Flask - 50 mL with ground glass stopper.

5.2   Balance - analytical, capable of accurately weighing ±0.0001g.
                                   D-63/VOA                            OLM03.0

-------
 Exhibit D — Volatiles Appendix A
 Screening of Hexadecane Extracts for  Volatiles


 5.3    Pyrex Glass Wool

 5.4    Balances  -  analytical,  capable  of  accurately weighing ±0.0001g,  and a
       top-loading balance  capable of  weighing 100 g  ±0.01g.   The  balances
       must be calibrated with class S weights or known  reference weights once
       per each  12-hour work shift.  The  balances must be  calibrated  with class
       S weights at a  minimum  of once  per month.  The balances  must also be
       annually  checked by  a certified technician.

 5.5    Centrifuge

 5.6    Gas Chromatograph/Mass  Spectrometer  (GC/MS) System

 5.6.1     Gas Chromatograph -  an analytical system complete  with a  temperature
          programmable gas  Chromatograph  suitable for on-column injection and
          all required accessories including syringes, analytical columns,
          gases, detector and  strip-chart recorder.   A data  system  is
          recommended  for measuring peak  areas.

 5.6.2     Gas Chromatography Column -  30  m  (or longer) x 2 mm ID glass column
          packed with  10% OV-101  on 100-120 mesh chromosorb  W-HP (or
          equivalent).   The column temperature should be programmed from 80  °C
          to  280 °C  at  16 C°/min.  and  held at 280 °C  for 10  minutes.

 5.6.3     Flame  lonization  Detector

 6.0   REAGENTS AND  STANDARDS

 6.1   Reagents

 6.1.1     Reagent Water  - defined  as water in which an interferent  is not
          observed at  the CRQL of each analyte of interest.

 6.1.2     Hexadecane and Methanol - pesticide residue analysis  grade or
          equivalent.

 6.2   Standards

 6.2.1     Introduction

          The Contractor must provide all standard solutions to be  used  with
         this contract.  These standards may be used only after they have been
          certified according to the procedure in Exhibit E.  The Contractor
         must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained  by the
          contractor and presented upon request.

6.2.2    Stock Standard Solutions

6.2.2.1     Stock standard solutions  (1 /xg//iL)  can be prepared from pure
             standard materials or purchased as certified solutions.
                                   D-64/VOA                            OLM03.0

-------
                                             Exhibit D — Volatiles Appendix A
                                Screening of Hexadecane Extracts for Volatiles
6.2.2.2     Prepare stock standard solutions by accurately weighing about
            0.01 g of pure material.  Dissolve the material in methanol and
            dilute to volume in a 10 mL volumetric flask.  Larger volumes can
            be used at the convenience of the analyst.  If compound purity,is
            certified at 97% or greater, the weight can be used without
            correction to calculate the concentration of the stock standard.

6.2.3    Working Standard Solutions

6.2.3.1     Standard Mixture #1

            Prepare a working standard mixture containing benzene, toluene,
            ethylbenzene and xylene at 100 ng//xL of each compound in methanol.

6.2.3.2     Standard Mixture #2

            Prepare a working standard mixture containing n-nonane and n-
            dodecane at 100 ng//tL of each compound in methanol.

6.2.4    Storage of Standards

         Transfer all standard solutions into multiple Teflon-sealed screw-cap
         vials.  Store, with no head-space, at -10 °C to -20 °C,  and protect
         from light.  Stock standard solutions should be checked frequently
         for signs of degradation or evaporation,  especially just prior to
         preparing calibration standards from them.  These solutions must be
         replaced after six months,  or sooner,  if comparison with quality
         control check samples indicates a problem.  Standards prepared from
         gases or reactive compounds such as styrene must be replaced after
         two months, or sooner, if comparison with quality control check
         samples indicates a problem.

7.0   QUALITY CONTROL

7.1    Method Blank

7.1.1    Summary

         A  method blank is a volume of clean reagent water taken through the
         extraction and screening procedure.  The volume of reagent water used
         must be approximately equal to the volume of associated samples.  The
         purpose of a  method blank is to determine the levels of
         contamination associated with the processing and analysis of samples.

7.1.2    Frequency

         One  method blank must be extracted and analyzed on each GC/FID
         system used to screen samples for the following,  whichever is most
         frequent.

         •   Each SDG,  or

         •   Each 20 samples in a SDG, including matrix spike and matrix spike
             duplicate, or

                                   D-65/VOA                            OLM03.0

-------
 Exhibit  D  —  Volatiles Appendix A
 Screening  of  Hexadecane Extracts for Volatiles
          •  When  samples  are extracted.

 7.1.3     Procedure

          For screening of  volatile organics, a method blank consists of  a  40
          mL volume of reagent water extracted with 2 itiL of hexadecane.   The
          hexadecane extract is then screened on a GC/FID system.

 8.0   CALIBRATION  AND STANDARDIZATION

 8.1   GC/FID Operating Conditions

      Refer to Section 5.5.2 for recommended column temperature program.

 8.2   GC  Calibration

 8.2.1     Summary

          Prior to  sample analysis each GC/FID system must be standardized  for
          half scale response.

 8.2.2     Frequency

          Each GC/FID system must be calibrated at the beginning of each  12-
          hour shift.

 8.2.3     Procedure

 8.2.3.1     Add 200 \iL of each of working standard mixtures #1 and #2  (Section
            6.2.3) to separate 40 mL portions of reagent water in 50 mL
            volumetric flasks.   Immediately add 2 mL of hexadecane, cap  the
            flask, and shake vigorously for 1 minute.  Let phases separate.
            Open the flask and add sufficient reagent water to bring the
            hexadecane layer into the neck of the flask.  Transfer
            approximately 1 mL of the hexadecane layer to a 2 mL GC vial.

8.2.3.2     Inject 1-2 /xL of the extracts that contain approximately 10  ng//*L
            each of standard mixture #1 and standard mixture #2 compounds.

9.0   PROCEDURE

9.1   Sample Preparation

9.1.1    Water

9.1.1.1     Allow the contents of the 40 mL sample vial to come to room
            temperature.   Quickly transfer the contents of the 40 mL sample
            vial to a 50 mL volumetric flask.  Immediately add 2 mL of
            hexadecane,  cap the flask, and shake vigorously for 1 minute.  Let
            phases separate.  Open the flask and add sufficient reagent  water
            to bring the hexadecane layer into the neck of the flask.
                                   D-66/VOA                            OLM03.0

-------
                                             Exhibit D — Volatiles Appendix A
                                Screening of Hexadecane Extracts for Volatiles
9.1.1.2     Transfer approximately 1 tnL of the hexadecane layer to a 2 mL GC
            vial.  If an emulsion is present after shaking the sample, break
            it by doing the following:

            •   Pulling the emulsion through a small plug of Pyrex glass wool
                packed in a pipet, or

            •   Transferring the emulsion to a centrifuge tube and
                centrifuging for several minutes.

9.1.2    Soil/Sediment

         Add approximately 10 g of soil/sediment (wet weight) to 40 mL of
         reagent water in a 50 mL centrifuge tube with a ground glass stopper
         or Teflon-lined cap.  Cap and shake vigorously for 1 minute.
         Centrifuge the capped flask briefly.  Quickly transfer supernatant
         water to a 50 mL volumetric flask equipped with a ground-glass
         stopper.  Follow 9.1.1 starting with the addition of 2 mL of
         hexadecane.

9.2   GC/FID Analysis

      Inject the same volume of sample hexadecane extract as the extracted
      standard mixture in 8.2.3.

9.2.1    GC/FID chromatogram Interpretation — Following are two options for
         interpreting the GC/FID Chromatograms.

9.2.1.1     Option A is to use standard mixture #1 containing the aromatics to
            calculate an approximate concentration of the aromatics in the
            sample.  Use this information to determine the proper dilution for
            purge and trap if the sample is water, or whether to use the low
            or medium level GC/MS purge and trap methods if the sample is
            soil/sediment (see Table 1, Section 9.3 for guidance).  This
            should be the best approach; however, the aromatics may be absent
            or obscured by higher concentrations of other purgeables.  In
            these cases, Option B may be the best approach.

9.2.1.2     Option B is to use standard mixture #2 containing n-nonane and n-
            dodecane to calculate a factor.  Use the factor to calculate a
            dilution for purge and trap of a water sample or to determine
            whether to use the low or medium level GC/MS purge and trap
            methods for soil/sediment samples (see Table 1, Section 9.3 for
            guidance).  All purgeables of interest have retention times less
            than the n-dodecane.

9.3   Analytical Decision Point

9.3.1    Water

9.3.1.1     Compare the chromatograms of the hexadecane extract of the sample
            with those of the reagent blank and extract of the standard.
                                   D-67/VOA                            OLM03.0

-------
Exhibit D — Volatiles Appendix A
Screening of Hexadecane Extracts for Volatiles
9.3.1.2     If no peaks are noted, other than those also in the reagent blank,
            analyze a 5 mL water sample by purge and trap GC/MS.

9.3.1.3     If peaks are present prior to the n-dodecane and the aromatics are
            distinguishable, follow Option A (Section 9.2.1).

9.3.1.4     If peaks are present prior to the n-dodecane but ths aromatics are
            absent or indistinguishable, use option B as follows:  if all
            peaks are ^3% of the n-nonane, analyze a 5 mL water sample by
            purge and trap GC/MS.  If any peaks are S3%of the n-nonane,
            measure the peak height or area of the major peak and calculate
            the dilution factor as follows:

            Peak area of sample major peak   x  50  =  dilution factor
                Peak area of n-nonans

            The water sample will be diluted using the calculated factor just
            prior to purge and trap GC/MS analysis.

9.3.2    Soil/Sediment

9.3.2.1     Compare the chroraatograms of the hexadecane extract of the sample
            with those of the reagent blank and extract of the standard.

9.3.2.2     If no peaks are noted, other than those also in the reagent blank,
            analyze a 5 g sample by low level GC/MS.

9.3.2.3     If peaks are present prior to the n-dodecane and the aromatics are
            distinguishable, follow Option A (Section 9.2.1) and the
            concentration information in Table 1,  to determine whether to
            analyze by low or medium level method.

9.3.2.4     If peaks are present prior to the n-nonane but the aromatics are
            absent or indistinguishable, and using Option B as follows,
            calculate a factor using the following formula:

            Peak area of sample major peak   =  X Factor
               Peak area of n-nonane

                                   Table  1
                 Determination of GC/MS Purge and Trap Method
                                               Approximate Concentration
     X Factor	Analyze by	Range* (^g/Kg)	

      0-1.0          low level method          0-1,200
      > l.o          medium level method       >1,200

               *  This concentration range is based on the response of
               aromatics to GC/FID.   When comparing GC/FID responses, the
               concentration for halomethanes is 20 times higher, and that for
               haloethanes is 10 times higher.

                                   D-68/VOA                            OLM03.0

-------
     EXHIBIT D
ANALYTICAL METHODS
 FOR  SEMIVOLATILES
     D-l/SVOA                            OLM03.0

-------
               Exhibit D - Analytical Methods for Semivolatiles

                               Table of  Contents


Section                                                                   Page


1.0   SCOPE AND APPLICATION 	    4

2.0   SUMMARY OF METHOD	    5
      2.1   Water	    5
      2.2   Low Soil/Sediment	    5
      2.3   Medium Soil/Sediment  	    5

3.0   DEFINITIONS 	    5

4.0   INTERFERENCES 	    6

5.0   SAFETY	    6

6.0   EQUIPMENT AND SUPPLIES  	    7

7.0   REAGENTS AND STANDARDS	   13
      7.1   Reagents	   13
      7.2   Standards	   13

7.3   Storage of Standard Solutions 	   16

8.0   SAMPLE COLLECTION,  PRESERVATION AND STORAGE 	   18
      8.1   Sample Collection and Preservation  	   18
      8.2   Procedure for Sample Storage  	   18
      8.3   Procedure for Sample Extract Storage  	   18
      8.4   Contract Required Holding Times 	   18

9.0   CALIBRATION AND STANDARDIZATION 	   19
      9.1   Instrument Operating Conditions 	   19
      9.2   GC/MS Mass Calibration  (Tuning) and Ion Abundance 	   19
      9.3   Initial Calibration 	   21
      9.4   Continuing Calibration  	   23

10.0  PROCEDURE	   27
      10.1  Sample Preparation	   27
      10.2  Concentrating the Extract	   32
      10.3  Sample Cleanup by GPC	   34
      10.4  Sample Extract Cleanup by GPC	   39
      10.5  Final Concentration 	   41
      10.6  Sample Analysis by GC/MS	   41

11.0  DATA ANALYSIS AND CALCULATIONS	   43
      11.1  Qualitative Identification  	   43
      11.2  Calculations	   46
      11.3  Technical Acceptance Criteria for Sample Analysis 	   50
      11.4  Corrective Action 	   50


                                   D-2/SVOA                            OLM03.0

-------
12.0  QUALITY CONTROL	    54
      12.1  Method Blanks	    54
      12.2  Matrix Spike/Matrix Spike Duplicate  (MS/MSD)   	    56

13.0  METHOD PERFORMANCE   	    60

14.0  POLLUTION PREVENTION   	    60

15.0  WASTE MANAGEMENT  	    60

16.0  REFERENCES	    60

17.0  TABLES/DIAGRAMS/FLOWCHARTS   	    61

APPENDIX A -  SCREENING OF SEMIVOLATILE ORGANIC EXTRACTS   	    70
      1.0   SCOPE AND APPLICATION	    70
      2.0   SUMMARY OF METHOD	    70
      3.0   INTERFERENCES	    71
      4.0   SAFETY	    71
      5.0   EQUIPMENT AND SUPPLIES	    72
      6.0   REAGENTS AND STANDARDS	    74
      7.0   QUALITY CONTROL	    75
      8.0   CALIBRATION AND  STANDARDIZATION  	    76
      9.0   PROCEDURE	    77
                                   D-3/SVOA                             OLM03.0

-------
Exhibit D Semivolatiles — Section 1
Scope and Application
1.0   SCOPE AND APPLICATION

1.1   The analytical method that follows is designed to analyze water, soil
      and sediment from hazardous waste sites for the semivolatile organic
      compounds on the Target Compound List (TCL), (see Exhibit C).

1.2   The method is based on EPA Method 625 (Base/Neutrals and Acids) and it
      covers the determination of a number of organic compounds that are
      partitioned into an organic solvent and are amenable to gas
      chromatography.   These target compounds and the contract required
      quantitation limits are listed in Exhibit C.

1.3   The method involves solvent extraction of the matrix sample,
      characterization to determine the appropriate analytical protocol to be
      used followed by appropriate cleanup procedure and GC/MS analysis to
      determine the semivolatile organic compounds present in the sample.

1.4   Problems have been associated with the following compounds analyzed by
      this method:

1.4.1    Dichlorobenzidine and 4-chloroaniline can be subject to oxidative
         losses during solvent concentration.

1.4.2    Hexachlorocyclopentadiene is subject to thermal decomposition in the
         inlet of the gas chromatograph, chemical reactions in acetone
         solution, and photochemical decomposition.

1.4.3    N-nitrosodiphenylamine decomposes in the gas chromatograph inlet
         forming diphenylamine and, consequently, may be detected as
         diphenylamine.
                                   D-4/SVOA                            OLM03.0

-------
                                     Exhibit D Semivolatiles — Sections 2 & 3
                                                 Summary of Method/Definitions
 2.0    SUMMARY OF METHOD

 2.1    Water

       A one liter aliquot of sample  is  acidified to pH 2.0 and extracted with
       tr.ethylene chloride using a continuous liquid-liquid extractor.
       Separatory funnel extraction is NOT permitted.  The methylene chloride
       extract is dried with sodium sulfate, concentrated, subjected to GPC
       (GPC is required when higher molecular weight compounds are present that
       interfere with the analyses of target compounds; GPC is optional for all
       other circumstances), and analyzed by GC/MS for extractable organics.

 2.2    Low Soil/Sediment

       A thirty (30) gram portion of soil/sediment is mixed with anhydrous
       powdered sodium sulfate and extracted with 1:1 methylene
       chloride/acetone solution using an ultrasonic probe. If the low level
       screen (Appendix A) is used, a portion of this dilute extract is
       concentrated fivefold and screened by GC/FID or GC/MS.  If peaks are
       present at greater than 10,000 /xg/kg, discard the extract and prepare
       the sample by the medium level method.  If no peaks are present at
       greater than 10,000 ^g/kg the entire extract is concentrated, subjected
       to GPC cleanup, and analyzed by GC/MS for extractable organics.

2.3   Medium Soil/Sediment

      Approximately one gram portion of soil/sediment is mixed with anhydrous
      powdered sodium sulfate in a vial and extracted with methylene chloride.
      The methylene chloride extract can be screened for extractable organics
      by GC/FID or GC/MS.  A method for screening is found in Appendix A.  If
      organic compounds are detected by the screen,  the methylene chloride
      extract is subjected to GPC cleanup and analyzed by GC/MS for
      extractable organics.   If no organic compounds are detected by the
      medium level screen,  then a low level sample preparation is required.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                   D-5/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Sections 4 & 5
Interferences/Safety
4.0   INTERFERENCES

      Contaminants in solvents, reagents,  glassware,  and other sample
      processing hardware may cause method interferences such as discrete
      artifacts and/or elevated baselines  in the extracted ion current
      profiles (EICPs).  All of these materials routinely must be demonstrated
      to be free from interferences under  the conditions of the analysis by
      running laboratory method blanks.  Matrix interferences may be caused by
      contaminants that are coextracted from the sample. The extent of matrix
      interferences will vary considerably from source to source.

5.0   SAFETY

      The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely determined; however, each chemical should be treated
      as a potential health hazard.  Exposure to these reagents should be
      reduced to the lowest possible level.   The laboratory is responsible for
      maintaining a current awareness file of OSHA regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
      file of data handling sheets should  be made available to all personnel
      involved in these analyses.   Specifically,  concentrated sulfuric acid
      presents some hazards and is moderately toxic and extremely irritating
      to skin and mucous membranes.  Use these reagents in a fume hood
      whenever possible and if eye or skin contact occurs flush with large
      volumes of water.  Always wear safety glasses or a shield for eye
      protection,  protective clothing,  and observe proper mixing when working
      with these reagents.
                                   D-6/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 6
                                                        Equipment and Supplies
6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, catalog and part numbers are for illustrative
      purposes only.  No endorsement is implied.  Equivalent performance may
      be achieved using equipment and supplies other than those specified
      here, but demonstration of equivalent performance meeting the
      requirements of this Statement of Work is the responsibility of the
      Contractor.  The Contractor shall document any use of alternate
      equipment or supplies in the SDG Narrative.

6.1   Glassware

6.1.1    Continuous Liquid-Liquid Extractors - equipped with Teflon or glass
         connecting joints and stopcocks requiring no lubrication (Hershberg-
         Wolf extractor. Ace Glass Company, Vineland, NJ P/N 6841-10 or
         equivalent) or Hydrophobic Membrane-based Extractor (Accelerated One
         Step™ Extractor, Corning series 3195 or equivalent).

6.1.2    Beakers - 400 inL.

6.1.3    Syringes - 2 /xL, 10 /xL, 0.2 mL, 0.5 mL and 10 mL with Luerlok
         fitting.

6.1.4    Glass Scintillation Vials - at least 20 mL, with screw-cap and Teflon
         or aluminum foil liner.

6.1.5    Pasteur Pipets - 1 mL glass, disposable.

6.1.6    Vial and Caps - amber glass, 2 mL capacity with Teflon-lined screw
         cap, 2 mL capacity for GC auto sampler.

6.1.7    Drying Column - 19 mm ID chromatographic column with coarse frit
         (substitution of a small pad of Pyrex glass wool for the frit will
         help prevent cross contamination of sample extracts).

6.2   Kuderna-Danish (K-D) Apparatus

6.2.1    Concentrator Tubes - 15 mL and 10 mL graduated (Kontes K-570050-1025
         or K-570040-1025 or equivalent).  Calibration must be checked at the
         volumes employed in the test.  Ground-glass stoppers are used to
         prevent evaporation of extracts.

6.2.2    Evaporative Flasks - 500 mL (Kontes K-570001-0500 or equivalent).
         Attach to concentrator tube with springs.

6.2.3    Snyder Column - three-ball macro (Kontes K-503000-0121 or
         equivalent).

6.2.4    Snyder Column - two-ball micro (Kontes K-569001-0219 or equivalent).

6.3   Spatula - stainless steel or Teflon
                                   D-7/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 6
Equipment and Supplies
6.4   Balances - analytical, capable of accurately weighing ± 0.0001 g and one
      capable of weighing 100 g to ±0.01 g.  The balances must be calibrated
      with class S weights or known reference weights once per each 12-hour
      work shift.  The balances must be calibrated with class S weights at a
      minimum of once per month.  The balances must also be annually checked
      by a certified technician.

6.5   Ultrasonic Cell Disrupters - Heat Systems,  Ultrasonics Inc., Model W-385
      Sonicator (475 watt with pulsing capability, No.200, 1/2 inch tapped
      disruptor horn, No. 419, 1/8 inch standard tapered Microtip probe, and
      No. 305, 3/4 inch tapped high gain "Q" disruptor horn, or No. 208 3/4
      inch standard solid disruptor horn),  or equivalent devices with a
      minimum of 375 watt output capability.  NOTE:  In order to ensure that
      sufficient energy is transferred to the sample during extraction, the
      Microtip probe or horn  shall be replaced if the tip begins to erode.
      Erosion of the tip is evidenced by a rough surface.

6.6   Sonabox Acoustic Enclosure (or equivalent)  - for use with disruptor to
      decrease noise level.

6.7   Vacuum Filtration Apparatus

6.7.1    Buchner Funnel

6.7.2    Filter Paper - Whatman No. 41 or equivalent

6.8   Pyrex Glass Wool - rinsed with methylene chloride

6.9   Test Tube Rack

6.10  Silicon Carbide Boiling Chips - approximately 10/40  mesh.  Heat to
      400 °C for 30 minutes or soxhlet extract with methylene chloride.
      Teflon Boiling chips solvent rinsed prior to use are acceptable.

6.11  Water Bath - heated, with concentric ring cover, capable of temperature
      control (±2 °C).   The bath should be used in a hood.

6.12  Oven - drying

6.13  Desiccator

6.14  Crucibles - porcelin

6.15  Nitrogen Evaporation Device - equipped with a water  bath that can be
      maintained at 35-40 °C.  (N-Evap by Organomation Associates, Inc., South
      Berlin, MA,  or equivalent).  To prevent the release  of solvent fumes
      into the laboratory, the nitrogen evaporator device  must be used in a
      hood.

6.16  pH Paper - including narrow range capable of measuring a pH of 2.0.

6.17  pH Meter - with a combination glass electrode, calibrate according to
      manufacturer's instructions.  The pH meter shall be  calibrated prior to
      each use.

                                   D-8/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 6
                                                        Equipment and Supplies
6.18  GPC Cleanup System

6.18.1   Gel Permeation Chromatography System - GPC Autoprep model 1002 A or
         B, Analytical Biochemical Laboratories, Inc., or equivalent.  Systems
         that perform satisfactorily have been assembled from the following
         components - an HPLC pump, an auto sampler or a valving system with
         sample loops, and a fraction collector.  All systems, whether
         automated or manual, must meet the calibration requirements of
         Section 10.3.3.

6.18.2   NOTE: GPC cleanup is required for all soil/sediment extracts, and for
         water extracts containing higher molecular weight contaminants that
         interfere with the analyses of the target compounds.

6.18.3   Chromatographic Column - 700 mm x 25 mm ID glass column.  Flow is
         upward.  To simplify switching from the UV detector during
         calibration to the GPC collection device during extract cleanup, an
         optional double 3-way valve (Rheodyne Type 50 Teflon Rotary Valve
         #10-262 or equivalent) may be attached so that the column exit flow
         can be shunted either to the UV flow-through cell or to the GPC
         collection device.

6.18.4   Guard Column (optional) - 5 cm,  with appropriate fittings to connect
         the inlet side of the analytical column (Supelco 5-8319 or
         equivalent).

6.18.5   Bio Beads (S-X3)  - 200-400 mesh, 70 g ^Bio-Rac Laboratories,
         Richmond, CA, Catalog 152-2750 or equivalent).  An additional 5 g of
         Bio Beads are required if the optional guard column is employed.  The
         quality of Bio Beads may vary from lot to lot because of excessive
         fines in some lots.   In addition to fines having a detrimental effect
         on chromatography, they can also pass through the column screens and
         damage the valve.

6.18.6   Ultraviolet Detector - fixed wavelength (254 nm) with a semi-prep
         flow-through cell.

6.18.7   Strip chart recorder, recording integrator or laboratory data system.

6.18.8   Syringe Filter Assembly, disposable - Bio-Rad "Prep Disc" sample
         filter assembly #343-0005, 25 mm, and 5 micron filter discs or
         equivalent.   Note:  Some instrument manufacturer's recommend a
         smaller micron size filter disc.  Consult your instrument operation
         manual to determine the proper filter disc to use in your system.
         Check each batch for contaminants.  Rinse each filter assembly (prior
         to use) with methylene chloride if necessary.
                                   D-9/SVOA                            OLM03.0

-------
 Exhibit D  Semivolatiles  —  Section 6
 Equipment  and Supplies
 6.19  Gas Chromatograph/Mass Spectrometer  (GC/MS) System

 6.19.1   Gas Chromatograph - The gas chromatograph (GC) system must be capable
         of temperature programming and have a flow controller that maintains
         a constant column flow rate throughout the temperature program
         operations.  The system must b~ suitable for splitless injection and
         have all required accessories including syringes, analytical columns,
         and gases.  All GC carrier gas lines must be constructed: from
         stainless steel or copper tubing.  Non-polytetrafluoroethylene  (PTFE)
         thread sealants or flow controllers with rubber components are  not to
         be used.

 6.19.2   Gas Chromatography Column - Minimum length 30 m x 0.25 mm ID (or 0.32
         mm) bonded-phase silicon coated fused silica capillary column DB-5
         (JSW Scientific); RTx-5 (Restek); SPB-5 (Supelco); AT-5 (Alltech);
         HP-5 (Hewlett-Packard); CP-Sil 8CB (Chrompack); 007-2 (Quadrex); BP-5
         (SGE); or equivalent.  Note that this is a minimum requirement  for
         column length.  Longer columns may be used.   Although a film
         thickness of 1.0 micron is recommended because of its larger
         capacity, a film thickness of 0.25 micron may be used.  A description
         of the GC column used for analysis shall be provided in the SDG
         narrative.

 6.19.2.1    A capillary column is considered equivalent if:

            •   The column does not introduce contaminants which interfere
                with the identification and quantitation of the compounds
                listed in Exhibit C (Semivolatiles).

            •   The analytical results generated using the column meet the
                initial and continuing calibration technical acceptance
                criteria listed in the SOW, and the CRQLs listed in Exhibit C
                (Semivolatiles).

            •   The column can accept up to 160 ng of each compound listed in
                Exhibit C (Semivolatiles)  without becoming overloaded.

            •   The column provides equal or better resolution of the
                compounds listed in Exhibit C (Semivolatiles)  than the columns
                listed in Section 6.19.2.

6.19.2.2    As applicable,  follow manufacturer's instructions for use of its
            product.

6.19.2.3    The Contractor must maintain documentation that the column met the
            criteria in Section 6.19.2.1.   The minimum documentation is as
            follows:

6.19.2.3.1      Manufacturer provided information concerning the performance
               characteristics of the column;

6.19.2.3.2      Reconstructed ion chromatograms and data system reports
               generated on the GC/MS used for CLP analyses:
                                  D-10/SVOA                            OLM03.1

-------
                                          Exhibit D Semivolatiles — Section 6
                                                        Equipment and Supplies
                •    From blanks which demonstrate that there are no
                    contaminants which  interfere with the  semivolatile analysis
                    when using the column;

                •    For initial calibration standards analyzed using the
                    column;

                •    For continuing calibration standards analyzed using the
                    column.

6.19.2.3.3      Based on the Contractor generated data described in 6.19.2.3.2,
                the  Contractor must complete a written review, signed by the
                Laboratory Manager certifying that:

                •    The column performance meets the technical acceptance
                    criteria in 9.3.5 and 9.4.5.

                •    The low point initial calibration standard analysis has
                    adequate sensitivity to meet the semivolatile CRQLs.

                •    The high point initial calibration standard analysis was
                    not overloaded.

                •    The column does not introduce contaminants which interfere
                    with the identification and/or quantitation of compounds
                    listed in Exhibit C (Semivolatiles).

6.19.2.4    The documentation must be made available to the Agency during on-
            site laboratory evaluations or sent to the Agency upon request of
            the Technical Project Officer or the Administrative Project
            Officer.

6.19.2.5    Packed  columns cannot be used.

6.19.3   Mass Spectrometer - must be capable of scanning from 35 to 500 amu
         every  1 second or less,  utilizing 70 volts (nominal) electron energy
         in the electron impact ionization mode and producing a mass spectrum
         which meets the tuning acceptance criteria when 50 ng of
         decafluorotriphenylphosphine (DFTPP) is injected through the GC
         inlet.  The instrument must be vented to the outside of the facility
         or to a trapping system which prevents the release of contaminants in
         to the instrument room.

6.19.4   GC/MS  interface - any gas chromatograph to mass spectrometer
         interface  which provides acceptable sensitivity at contract required
         quantitation limits.   Gas chromatograph to mass spectrometer
         interfaces constructed of all-glass or glass-lined materials are
         recommended.  Glass can be deactivated by silanizing with
         dichlorodimethylsilane.

6.19.5   Data system - a computer system must be interfaced to the mass
         spectrometer that allows the continuous acquisition and storage on
         machine-readable media,  of all mass spectra obtained throughout the
         duration of the chromatographic program.   The computer must have

                                  D-11/SVOA                            OLM03.1

-------
Exhibit D Semivolatiles — Section 6
Equipment and Supplies
         software that allows searching any GC/MS data file for ions of a
         specified mass and plotting such ion abundance versus time or scan
         number.  This type of plot is defined as an Extracted Ion Current
         Profile (EICP).  Software must also be available that allows
         integrating the abundance in any EICP between specified time or scan
         number limits.  Also, for the non-target compounds, software must be
         available that allows for the comparison of sample spectra against
         reference library spectra.  The NIST/EPA/NIH (May 1992 release or
         later) and/or Wiley (1991 release or later), or equivalent mass
         spectral library shall be used as the reference library.  The
         operational data system must be able to flag all data files that have
         been edited manually by laboratory personnel.

6.19.6   Magnetic tape storage device - must be capable of recording data and
         suitable for long-term, off-line storage of GC/MS data.
                                   D-12/SVOA                           OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 7
                                                        Reagents and Standards
7.0   REAGENTS AND STANDARDS

7.1   Reagents

7.1.1    Reagent Water - defined as water in which an interferant is not
         observed at or above the CRQL for each analyte of interest.  Reagent
         water may be generated by passing tap water through a carbon filter
         bed containing about 453 g of activated carbon (Calgon Corp.,
         Filtrasorb - 300 or equivalent).

7.1.2    Sodium Thiosulfate - (ACS) granular.

7.1.3    Sulfuric Acid Solution (H2SO4)  - (1+1)  slowly  add  50 mL  of
         concentrated H2S04  (sp. gr.  1.84;  18 N) to  50  mL of reagent water.

7.1.4    Acetone, methanol,  methylene chloride, iso-octane, 2-propanol, and
         toluene - pesticide residue analysis grade or equivalent.

7.1.5    Sodium Sulfate - powdered or granular anhydrous reagent grade, heated
         at 400 °C for four hours in a shallow tray, cooled in a desiccator
         and stored in a glass bottle (Baker anhydrous powder,  catalog #73898;
         Baker anhydrous granulated, catalog #3375; or equivalent).   CAUTION:
         An open container of sodium sulfate may become contaminated during
         storage in the laboratory.

7.1.6    Sodium Sulfite - reagent grade.

7.2   Standards

7.2.1    Introduction

         The Contractor must provide all standards to be used with this
         contract.  These standards may be used only after they have been
         certified according to the procedure in Exhibit E. The Contractor
         must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

7.2.2    Stock Standard Solutions

7.2.2.1     Stock standard solution may be purchased or prepared using the
            following procedure.

7.2.2.1.1      Prepare stock standard solutions by accurately weighing about
               0.0100 g of pure material.  Dissolve the material in methylene
               chloride or another suitable  solvent and dilute to volume in a
               10 mL volumetric flask.  Larger volumes may be used  at the
               convenience of the analyst.

7.2.2.1.2      When compound purity is assayed to be 97.0 percent or greater,
               the weight may be used without correction to calculate the
               concentration of the stock solution.  If the compound purity is
               assayed to be less than 97.0  percent, the weight must be
                                   D-13/SVOA                           OLM03.0

-------
Exhibit D  Semivolatiles  — Section 7
Reagents and Standards
               corrected when calculating the concentration of the  stock
               solution.  See Exhibit E  (Analytical Standards Requirements).

7.2.2.1.3      Fresh stock standards must be prepared once every twelve
               months, or sooner if standards have degraded or concentrated.
               Stock standards must be checked for signs of degradation or
               concentration just prior to preparing secondary dilution and
               working standards from them.

7.2.3    Secondary Dilution Standards

7.2.3.1     Using stock standards, prepare secondary dilution standards in
            methylene chloride that contain the compounds of interest either
            singly or mixed together.

7.2.3.2     Fresh secondary dilution standards must be prepared once every
            twelve months, or sooner if standards have degraded or
            concentrated.  Secondary dilution standards must be checked for
            signs of degradation or concentration just prior to preparing
            working standards from them.

7.2.4    Working Standards

7.2.4.1     Surrogate Standard Spiking Solution

            Prepare a surrogate standard spiking solution that contains
            nitrobenzenene-dS,  terphenyl-d!4,  2-fluorobiphenyl,  and 1,2-
            dici.lorobenzene-d4 at a concentration of 100 /tg/mL;  phenol-d5,
            2,4,6-tribromophenol, 2-fluorophenol and 2-chlorophenol-d4 at a
            concentration of 150 /xg/mL.   Surrogate standards are added to all
            samples and calibration solutions.   Additional surrogates may be
            added at the laboratory's discretion.

7.2.4.2     Matrix Spiking Solution

7.2.4.2.1      The matrix spiking solution consists of the following:

               Bases/Neutrals                  Acids

               1,2,4-Trichlorobenzene         Pentachlorophenol
               Acenaphthene                   Phenol
               2,4-Dinitrotoluene             2-Chlorophenol
               Pyrene                          4-Chloro-3-methylphenol
               N-Nitroso-di-n-propylamine     4-Nitrophenol
               1,4-Dichlorobenzene
7.2.4.2.2      Prepare a spiking solution that contains each of the
               base/neutral compounds above at 100 ng/mL in methanol and the
               acid compounds at 150 /ug/mL in methanol.
                                  D-14/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 7
                                                        Reagents and Standards
7.2.4.3     GPC Calibration Solution

7.2.4.3.1      Prepare a calibration solution in methylene chloride containing
               the following analytes at the minimum concentrations listed (in
               elution order):
                Compound                       Concentration (mq/mL)

                Corn oil                                  25.0
                bis(2-ethylhexyl)phthalate                 0.5
                Methoxychlor                               0.1
                Perylene                                   0.02
                Sulfur (optional)                          0.08
7.2.4.3.2      NOTE:  Sulfur is not very soluble in methylene chloride, but it
               is soluble in warm corn oil.  Therefore, one approach is to
               weigh out the corn oil, warm it, and transfer the weighed
               amount of sulfur into the warm corn oil.  Mix it and then
               transfer into a volumetric flask with methylene chloride, along
               with the other calibration compounds.

7.2.4.4     Instrument Performance Check Solution.

            Prepare a solution of decafluorotriphenylphosphine (DFTPP), such
            that a 2 jxL injection will contain 50 ng of DFTPP.  The DFTPP may
            also be included in the calibration standards at this level.

7.2.4.5     Initial and Continuing Calibration Solutions.

7.2.4.5.1      Prepare calibration standards at a minimum of five
               concentration levels (20, 50, 80, 120, and 160 total ng per 2
               jiL).  Each calibration standard should contain each compound of
               interest and each surrogate.  Eight compounds (2,4-
               Dinitrophenol, 2,4,5-Trichlorophenol, 2-Nitroaniline, 3-
               Nitroaniline, 4-Nitroaniline, 4-Nitrophenol, 4,6-Dinitro-2-
               methylphenol, and Pentachlorophenol) will require only a four-
               point initial calibration at 50, 80, 120, and 160 total ng per
               2 /iL, since detection at less than 50 ng per injection is
               difficult.

7.2.4.5.2      In order to facilitate the confirmation of single component
               pesticides from the semivolatile library search data (see
               Exhibit D (Pesticides/Aroclors), Section 12.1.2),  the
               laboratory may include the single component pesticide target
               compounds listed in Exhibit C in the semivolatile continuing
               calibration standard.  The laboratory may add any or all of
               these compounds to the semivolatile continuing calibration
               standard, but at a concentration of 10 ng//*L or less.  Do not
               include the Aroclors or toxaphene mixtures in the semivolatile
               initial and continuing calibration standards.  If added to this
               GC/MS standard, these additional analytes are not reported on
               the semivolatile calibration form (Form VII), but must be
               included in the quantitation report for the continuing

                                   D-15/SVOA                           OLM03.0

-------
Exhibit D Semivolatiles — Section 7
Reagents and Standards


               calibration standard.  As only a single point calibration would
               be performed,  no %RSD or percent difference criteria would
               apply to these additional analytes.

7.2.4.5.3      The 50 ng/2 fiL calibration standard  is the continuing
               calibration standard.

7.2.4.6     Internal Standard Solution.

            An internal standard solution can be prepared by dissolving 100 mg
            of each of the following compounds in 50 mL of methylene chloride:
            l,4-dichlorobenzene-d4,  naphthalene-d8,  acenaphthene-dlO,
            phenanthrene-dlO, chrysene-d!2 and perylene-d!2.  It may be
            necessary to use 5.0 to 10.0 percent toluene in this solution and
            a few minutes of ultrasonic mixing in order to dissolve all the
            constituents.  The resulting solution will contain each standard
            at a concentration of 2000 ng//iL.  A 10 /±L portion of this
            solution should be added to each 1 mL of sample extract just prior
            to analysis by GC/MS.  This will result in 40 ng of each internal
            standard in the 2 ^L volume of extract  injected into the GC/MS.

            Note: For automated systems using an injection volume of less than
            10 jtL, the internal standard solution may need to be prepared at a
            different concentration.  Prepare the internal standard solution
            such that the aliquot used by the system maintains the required 40
            ng of each internal standard in the 2 /*L volume of extract
            injected into the GC/MS.

7.2.5    Ampulated Standard Extracts

         Standard solutions purchased from a chemical supply house as
         ampulated extracts in glass vials may be retained for 2 years from
         the preparation date, unless the manufacturer recommends a shorter
         time period.  Standard solutions prepared by the Contractor which are
         immediately ampulated in glass vials may be retained for 2 years from
         the preparation date.  Upon breaking the glass seal, the expiration
         times listed in Sections 7.2.2.1.3, 7.2.3.2, and 7.3.3. will apply.
         The Contractor is responsible for assuring that the integrity of the
         standards have not degraded (see Section 7.3.5).

7.3   Storage of Standard Solutions

7.3.1    Store the stock and secondary dilution standard solutions at less
         than 4 °C but not greater than 6 °C in Teflon-lined screw-cap amber
         bottles. Fresh standards should be prepared every twelve months at a
         minimum.

7.3.2    Store the working standards at less than 4 °C but not greater than
         6 °C in Teflon-sealed containers.  The solution should be checked
         frequently  for stability.  These solutions must be replaced after
         twelve months or sooner  if comparison with quality control  check
         samples  indicates a problem.
                                   D-16/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 7
                                                        Reagents and Standards
7.3.3    The continuing calibration standard (50 ng) should be prepared weekly
         and stored at less than 4 °C but not greater than 6 °C.
         Refrigeration of the GPC calibration solution may cause the corn oil
         to precipitate.  Before use, allow the solution to stand at room
         temperature until the corn oil dissolves.  Replace this calibration
         solution every six months, or more frequently if necessary.
7.3.4    Protect all standards from light.  Samples, sample extracts and
         standards must be stored separately.

7.3.5    The Contractor is responsible for maintaining the integrity of
         standard solutions and verifying prior to use.  Storage of standard
         solutions in the freezer may cause some compounds to precipitate.
         This means at a minimum, the standards must be brought to room
         temperature prior to use,  checked for losses,  and checked that all
         components have remained in solution.   Additional steps may be
         necessary to ensure all components are in solution.
                                  D-17/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 7
Reagents and Standards
8.0   SAMPLE COLLECTION, PRESERVATION AND STORAGE

8.1   Sample Collection and Preservation

8.1.1    Water samples may be collected in 1 L (or 1 quart) amber glass
         containers, fitted with screw-caps lined with Teflon.  If amber
         containers are not available, the samples should be protected from
         light.  Soil samples may be collected in glass containers or closed
         end tubes  (e.g., brass sleeves) in sufficient quantity to perform the
         analysis.  The specific requirements for site sample collection are
         outlined by the Region.

8.1.2    All samples must be iced or refrigerated at 4 °C (±2 °C)  from the
         time of collection until extraction.

8.2   Procedure for Sample Storage

8.2.1    The samples must be protected from light and refrigerated at 4 °C (±2
         °C)  from the time of receipt until 60 days after delivery of a
         complete, reconciled data package to the Agency.  After 60 days the
         samples may be disposed of in a manner that complies with all
         applicable regulations.

8.2.2    The samples must be stored in an atmosphere demonstrated to be free
         of all potential contaminants.

8.3   Procedure for Sample Extract Storage

8.3.1    Sample extracts must be protected from light and stored at less than
         4 °C but not greater than 6 °C until 365 days after delivery of a
         reconciled, complete data package to the Agency.

8.3.2    Samples,  sample extracts,  and standards must be stored separately.

8.4   Contract Required Holding Times

8.4.1    Extraction of water samples by continuous liquid-liquid procedures
         shall be started within 5 days of Validated Time of Sample Receipt
         (VTSR).   Extraction of soil/sediment samples by sonication procedures
         shall be completed within 10 days of VTSR.  Note:  Separatory funnel
         extraction procedures are not permitted.

8.4.2    As part of the Agency's QA program,  the Agency may provide
         Performance Evaluation samples as standard extracts which the
         Contractor is required to prepare per the instructions provided by
         the Agency.  The extraction holding time (5 days after VTSR for water
         and 10 days after VTSR for soil/sediment) does not apply to
         Performance Evaluation samples received as standard extracts.

8.4.3    Extracts of water and soil/sediment samples must be analyzed within
         40 days following extraction.
                                   D-18/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 9
                                               Calibration and Standardization
            Instrument Operating Conditions/GC/MS Calibration and Ion Abundance
9.0   CALIBRATION AND  STANDARDIZATION

9.1   Instrument Operating Conditions

9.1.1    Gas Chromatograph

9.1.1.1     The following are the gas chromatographic analytical conditions.
            The conditions are recommended unless otherwise noted.

             Initial Column Temperature Hold      40 °C for 4 minutes
             Column Temperature Program           40-270 °C at 10 C°/min.
             Final Column Temperature Hold        270 °C;
                                                  Hold Required:  3 minutes
                                                  after all compounds listed
                                                  in Exhibit C (SVOA) have
                                                  eluted
             Injector Temperature                 250-300 °C
             Transfer Line Temperature            250-300 °C
             Source Temperature                   According to manufacturer's
                                                  specifications
             Injector                             Grob-type, splitless
             Sample Volume                        2  fj.L
             Carrier Gas                          Helium at 30 cm/sec

9.1.1.2     Optimize GC conditions for analyte separation and sensitivity.
            Once optimized, the same GC conditions must be used for the
            analysis of all standards, samples, blanks, matrix spikes and
            matrix spike duplicates.

9.1.2    Mass Spectrometer

         The following are the required mass spectrometer analytical
         conditions:

              Electron Energy   70 volts (nominal)
              Mass Range        35 to 500 amu
              Scan Time         Not to exceed 1 second per scan

9.2   GC/MS Mass Calibration (Tuning) and Ion Abundance

9.2.1    Summary of GC/MS Instrument Performance Check

         The GC/MS system must be tuned to meet the manufacturer's
         specifications, using a suitable calibrant such as perfluoro-tri-N-
         butylamine (FC-43)  or perfluorokerosene (PFK).  The mass calibration
         and resolution of the GC/MS system are verified by the analysis of
         the instrument performance check solution (Section 7.2.4.4).  Prior
         to the analysis of any samples; including matrix spike/matrix spike
         duplicates, blanks or calibration standards, the Contractor must
         establish that the GC/MS system meets the mass spectral ion abundance
                                  D-19/SVOA                            OLM03.0

-------
 Exhibit  D  Semivolatiles  —  Section 9
 Calibration  and  Standardization
 GC/MS  Mass Calibration and  Ion Abundance
          criteria  for the  instrument performance check solution containing
          decafluorotriphenylphosphine  (DFTPP).

 9.2.2     Frequency of GC/MS  Instrument Performance Check

 9.2.2.1     The instrument performance check  solution must be analyzed once at
            the beginning of each 12-hour period during which samples or
            standards are analyzed.

 9.2.2.2     The 12-hour time period for a GC/MS system instrument performance
            check and standards calibration (initial or continuing calibration
            criteria) begins at the moment of injection of the DFTPP analysis
            that the laboratory submits as documentation of a compliant
            instrument performance check.  The time period ends after 12 hours
            have elapsed according to the system clock.

 9.2.3     Procedure for GC/MS Instrument Performance Check

          The analysis of the instrument performance check solution may be
          performed as an injection of up to 50 ng of DFTPP into the GC/MS or
          by adding 50 ng of DFTPP to the calibration standards (Section
          7.2.4.5)  and analyzing the calibration standard.

 9.2.4     Technical Acceptance Criteria for GC/MS Instrument Performance Check

 9.2.4.1     The GC/MS system must be tuned at the frequency described in
            Section 9.2.2.

 9.2.4.2     The abundance criteria listed in Table 1 must be met for a 50 ng
            injection of DFTPP.  The mass spectrum of DFTPP must be acquired
            in the following manner:   Three scans (the peak apex scan and the
            scans  immediately preceding and following the apex)  are acquired
            and averaged.   Background subtraction is required,  and must be
            accomplished using a single scan acquired no more than 20 scans
            prior  to the elution of DFTPP.   Do not subtract part of the DFTPP
            peak.    Note:   All subsequent standards,  samples, MS/MSD, and
            blanks associated with a DFTPP analysis must use the identical
            mass spectrometer instrument conditions.

9.2.5    Corrective Action for GC/MS Instrument Performance Check

9.2.5.1     If the DFTPP acceptance criteria are not met, re-tune the GC/MS
            system.  It may be necessary to clean the ion source,  clean
            quadrupoles,  or take other actions to achieve the technical
            acceptance criteria.

9.2.5.2     DFTPP  acceptance criteria MUST be met before any standards,
            samples,  including MS/MSD,  or required blanks are analyzed.  Any
            standards,  samples, or required blanks analyzed when tuning •
            technical acceptance criteria have not been met will require
            reanalysis at  no additional cost to the Agency.
                                  D-20/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles  — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
 9.3    Initial Calibration

 9.3.1    Summary of  Initial Calibration

         Prior to the  analysis of  samples and required blanks, and after the
         instrument  performance check solution criteria have been met, each
         GC/MS system  must be calibrated at a minimum of five concentrations
         (Section 7.2.4.5) to determine instrument sensitivity and the
         linearity of  GC/MS response for the semivolatile target and surrogate
         compounds.

 9.3.2    Frequency of  Initial Calibration

 9.3.2.1     Each GC/MS system must be initially calibrated upon award of the
            contract, whenever the contractor takes corrective action which
            may change or affect the initial calibration criteria (e.g., ion
            source cleaning or repairs, column replacement, etc.), or if the
            continuing calibration technical acceptance criteria have not been
            met.

 9.3.2.2     If time  still remains  in the 12-hour time period after meeting the
            technical  acceptance criteria for the initial calibration, samples
            may be analyzed.  It is not necessary to analyze a continuing
            calibration standard within this 12-hour time period, if the
            initial  calibration standard that is the same concentration as the
            continuing calibration standard meets the continuing calibration
            technical acceptance criteria.  Quantitate all sample and quality
            control  sample results and quality control criteria results, such
            as internal standard area response change and retention time
            shift,  against the initial calibration standard that is the same
            concentration as the continuing calibration standard (50 ng/2 ^L).

9.3.3    Procedure for Initial Calibration

9.3.3.1     All standard/spiking solutions and blanks must be allowed to warm
            to ambient temperature (approximately 1 hour) before preparation
            or analysis.

9.3.3.2     Prepare  five calibration standards containing all the semivolatile
            target and surrogate compounds at the concentrations described in
            Section  7.2.4.5.

9.3.3.3     Add a 10 fiL, aliquot of internal standard solution (Section
            7.2.4.6)  to 1 mL aliquot of calibration standards to result in 40
            ng of internal standard in the 2 fiL volume of calibration standard
            injected onto the GC/MS.   The internal standards specified in
            Section  7.2.4.6 should permit most of the semivolatile target
            compounds to have relative retention times of 0.80 to 1.20, using
            the assignments of internal standards to target compounds given in
            Table 2.

9.3.3.4     Analyze  each calibration standard by injecting 2.0 /xL of standard.


                                   D-21/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 9
Calibration and Standardization
Initial Calibration
9.3.4    Calculations for Initial Calibration

9.3.4.1     Calculate the relative response factors (RRF) for each
            semivolatile target and surrogate compound using Equation 1 and
            the primary characteristic ions found in Table 3 (Internal
            Standards) and Table 4 (Target Compounds and Surrogates).  Assign
            the target compounds and surrogate to the internal standard
            according to Table 2.  For internal standards, use the primary ion
            listed in Table 3 unless interferences are present.  If
            interferences prevent the use of the primary ion for a given
            internal standards, use the secondary ion(s) listed in Table 3.

            NOTE: Unless otherwise stated, the area response of the primary
            characteristic ion is the quantitation ion.


            EQ' *                           A,    Cis
                                      RRF = 	fL  X —if
                                            Als    Cx

           Where,
           Ax =   Area of the characteristic ion for the compound to be
                 measured (see Table 4)
           AJS=   Area of the characteristic ion for specific internal
                 standard (see Table 3)
           Cjs=   Amount  of the internal  standard injected (ng)
           Cx =   Amount  of the compound  to be  measured injected (ng)

9.3.4.2     The mean relative response factor (RRF) must be calculated for all
            compounds.  Calculate the % Relative Standard Deviation  (%RSD) of
            the RRF values for the initial calibration using the following
            equation:
            EQ. 2
                        %RSD = Standard Deviation x  1QQ
                                      Mean
            Where,
             Standard Deviation  =
                                         (n-l)
            Xj =  each individual value used to calculate the mean
            x  =  the mean of n values
            n  =  the total number of values
                                   D-22/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles — Section 9
                                               Calibration and Standardization
                                                        Continuing Calibration
9.3.5    Technical Acceptance Criteria  for Initial Calibration

9.3.5.1     All initial calibration standards must be analyzed at the
            concentration  levels described in Section 7.2.4.5 and at the
            frequency described in Section 9.3.2 on a GC/MS system meeting the
            DFTPP technical acceptance  criteria.

9.3.5.2     The relative response factor  (RRF) at each calibration
            concentration  for each semivolatile target and surrogate compound
            must be greater than or equal to the compound's minimum acceptable
            relative response factor listed in Table 5.

9.3.5.3     The %RSD over  the initial calibration range for relative response
            factor for each semivolatile and surrogate compound that has a
            required %RSD  must be less  than or equal to the %RSD listed in
            Table 5.

9.3.5.4     Up to four compounds may fail to meet the criteria listed in Table
            5.  However, these four compounds must meet a minimum RRF
            criterion of 0.010 and have a %RSD less than or equal to 40.0
            percent.

9.3.5.5     Excluding those ions in the solvent front, no quantitation ion may
            saturate the detector, with the exception of quantitation ions in
            up to 3 compounds (including internal standards and surrogates)
            from the high  standard (160 ng/2 /*L).  Consu t the manufacturer's
            instrument manual to determine how saturation is indicated for
            your instrument.

9.3.6    Corrective Action for Initial Calibration

9.3.6.1     If any technical acceptance criteria for initial calibration are
            not met, inspect the system for problems.   It may be necessary to
            clean the ion  source,  change the column,  or take other corrective
            actions to achieve the acceptance criteria.

9.3.6.2     Initial calibration technical acceptance criteria must be met
            before any samples,  including matrix spike/matrix spike duplicates
            or required blanks are analyzed.   Any samples, including matrix
            spike/matrix spike duplicates or required blanks analyzed when
            initial calibration technical acceptance criteria have not been
            met will require reanalysis at no additional cost to the Agency.

9.4   Continuing Calibration

9.4.1    Summary of Continuing Calibration

         Prior to the analysis of  samples, including MS/MSD and required
         blanks,  and after tuning  criteria and initial calibration criteria
         have been met,  each GC/MS system must be routinely checked by
         analyzing a continuing calibration standard to ensure that the
         instrument continues to meet the instrument sensitivity and linearity
                                   D-23/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 9
Calibration and Standardization
Continuing Calibration
         requirements of the SOW.  The continuing calibration standard
         contains all the semivolatile target and surrogate compounds and
         internal standards.

9.4.2    Frequency of Continuing Calibration

9.4.2.1     Each GC/MS used for analysis must be calibrated once every 12-hour
            time period of operation.  The 12-hour time period begins with the
            injection of DFTPP.

9.4.2.2     If time still remains in the 12-hour time period after meeting the
            technical acceptance criteria for the initial calibration, samples
            may be analyzed.  It is not necessary to analyze a continuing
            calibration standard within this 12-hour time period, if the
            initial calibration standard that is the same concentration as the
            continuing calibration standard meets the continuing calibration
            technical acceptance criteria.  Quantitate all sample results
            against the initial calibration standard that is the same
            concentration as the continuing calibration standard (50 ng/2 ;xL).

9.4.3    Procedure for Continuing Calibration

9.4.3.1     All standard/spiking solutions and blanks must be allowed to warm
            to ambient temperature (approximately 1 hour) before preparation
            or analysis.

9.4.3.2     Add a 10 /xL aliquot of internal standard solution (paragraph
            7.2.4.6) to 1 mL aliquot of continuing calibration standard to
            result in 40 ng of internal standard in the 2 /xL volume of
            calibration standard injected onto the GC/MS.  The internal
            standards specified in paragraph 7.2.4.6 should permit most of the
            semivolatile target compounds to have relative retention times of
            0.80 to 1.20, using the assignments of internal standards to
            target compounds given in Table 2.

9.4.3.3     Analyze the continuing calibration standard by injecting 2.0 /u.L of
            standard.

9.4.4    Calculations for Continuing Calibration

9.4.4.1     Calculate a relative response factor (RRF) for each semivolatile
            target and surrogate compound using Equation 1 and the primary
            characteristic ions found in Table 3 (Internal Standards) and
            Table 4 (Target Compounds and Surrogates).  For internal
            standards, use the primary ions listed in Table 3 unless
            interferences are present.  If interferences prevent the use of
            the primary ion for a given internal standard, use the secondary
            ion(s) listed in Table 3.
                                   D-24/SVOA                            OLM03.0

-------
                                          Exhibit D Semivolatiles —  Section  9
                                               Calibration and  Standardization
                                                        Continuing Calibration
9.4.4.2     Calculate the  percent difference between the mean relative
            response factor  from the most recent  initial calibration  and the
            continuing  calibration  relative response factor for each
            semivolatile target and surrogate compound using Equation 3.

            EQ. 3
            Where:
                                        RRF. -  RRF,
                      % Difference,,,,,, =  		 x 100
            RRFj =      Mean relative response factor from the most recent
                        initial calibration meeting technical acceptance
                        criteria
            RRFC =      Relative response factor from continuing calibration
                        standard

9.4.5    Technical Acceptance Criteria for Continuing Calibration

9.4.5.1     The continuing calibration standard must be analyzed at the
            50 ng/2 ^L concentration level at the frequency described in
            Section 9.4.2, on a GC/MS system meeting the DFTPP tuning and the
            initial calibration technical acceptance ciiteria.

9.4.5.2     The relative response factor (RRF) for each semivolatile target
            and surrogate compound must be greater than or equal to the
            compound's minimum acceptable relative response factor listed in
            Table 5.

9.4.5.3     The relative response factor percent difference for each
            semivolatile target and surrogate compound that has a percent
            difference criteria must be within the inclusive range listed in
            Table 5.

9.4.5.4     Up to four semivolatile target compounds may fail to meet the
            minimum RRF or maximum percent difference criteria listed in Table
            5,  but the RRFs of those four compounds must be greater than or
            equal to 0.010, and the percent differences must be within the
            inclusive range of ±40.0 percent.

9.4.5.5     Excluding those ions in the solvent front, no quantitation ion may
            saturate the detector.   Consult the manufacturer's instrument
            operating manual to determine how saturation is indicated for your
            instrument.
                                   D-25/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 9
Calibration and Standardization
Continuing Calibration
9.4.6    Corrective Action for Continuing Calibration

9.4.6.1     If any continuing calibration technical acceptance criteria are
            not met, recalibrate the GC/MS instrument according to Section
            9.3.3.  It may be necessary to clean the ion source, change the
            column or take other corrective actions to achieve the continuing
            calibration technical acceptance criteria.

9.4.6.2     Continuing calibration technical acceptance criteria MUST be met
            before any samples,  including MS/MSD or required blanks are
            analyzed.  Any samples, including MS/MSD or required blanks
            analyzed when continuing calibration criteria have not been met
            will require reanalysis at no additional cost to the Agency.
                                  D-26/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                                            Sample Preparation
10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient sample amount (less than 90% of the required amount)
         is received to perform the analyses, the Contractor shall contact SMO
         to apprise them of the problem.   SMO will contact the Region for
         instructions.  The Region will either require that no sample analyses
         be performed or will require that a reduced volume be used for the
         sample analysis.  No other changes in the analyses will be permitted.
         The Contractor shall document the Region's decision in the SDG
         Narrative.

10.1.2   If multiphase samples (e.g., a two-phase liquid sample, oily
         sludge/sandy soil sample) are received by the Contractor, the
         Contractor shall contact SMO to apprise them of the type of sample
         received.   SMO will contact the Region.  If all phases of the sample
         are amenable to analysis, the Region may require the Contractor to do
         the following:

         •   Mix the sample and analyze an aliquot from the homogenized
             sample.

         •   Separate the phases of the sample and analyze each phase
             separately.  SMO will provide EPA sample numbers for the
             additional phases.

         •   Separate the phases and analyze one or more of the phases, but
             not all of the phases.  SMO will provide EPA sample numbers for
             the additional phases, if required.

         •   Do not analyze the sample.

10.1.2.1    If all of the phases are not amenable to analysis (i.e., .outside
            scope), the Region may require the Contractor to do the following:

            •   Separate the phases and analyze the phase(s) that is amenable
                to analysis.  SMO will provide EPA sample numbers for the
                additional phases, if required.

            •   Do not analyze the sample.

10.1.2.2    No other change in the analyses will be permitted.  The Contractor
            shall document the Region's decision in the SDG Narrative.

10.1.3   Water Samples

10.1.3.1    Continuous liquid-liquid extraction is used to extract the
            samples.  Separatory funnel extraction cannot be used.
                                   D-27/SVOA                           OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Sample Preparation


10.1.3.2    Continuous Liquid-Liquid Extraction Without Hydrophobic Membrane

10.1.3.2.1     Follow manufacturer's instructions for set-up.

10.1.3.2.2     Add methylene chloride to the bottom of the extractor and fill
               it to a depth of at least one inch above the bottom sidearm.

10.1.3.2.3     Measure out a 1 L sample aliquot in a separate,  clean graduated
               cylinder; transfer the aliquot to the continuous extractor.
               Adjust the pH to 2.0 with 1:1 H2SC>4 and verify it with a pH
               meter or narrow range pH paper.  Record the pH.   NOTE:  With
               some samples, it may be necessary to place a layer of glass
               wool between the methylene chloride and the water layer in the
               extractor to prevent precipitation of suspended solids into the
               methylene chloride during extraction.

10.1.3.2.4     Using a syringe or volumetric pipet,  add 0.5 mL of the
               surrogate standard spiking solution (7.2.4.1) into the sample
               and mix well.

10.1.3.2.5     Rinse the graduated cylinder with 50 mL of methylene chloride
               and transfer the rinsate to the continuous extractor.  If the
               sample was received in a 1 L container,  rinse the empty
               container with 50 mL of methylene chloride and add rinsate to
               the continuous extractor.

10.1.3.2.6     Add sufficient methylene chloride ;o the continuous extractor
               to ensure proper solvent cycling during operation.  Adjust the
               drip rate to 5 to 15 mL/minute (recommended); optimize the
               extraction drip rate.  Extract for a minimum of 18 hours.
               NOTE:  When a minimum drip rate of 10-15 mLs/min is maintained
               throughout the extraction, the extraction time may be reduced
               to a minimum of 12 hours.  Allow to cool, then detach the
               distillation flask.  Proceed to Section 10.2.

10.1.3.2.7     NOTE: Some continuous liquid-liquid extractors are also capable
               of concentrating the extract within the extraction set-up.
               Follow the manufacturer's instructions for concentration when
               using this type of extractor.

10.1.3.3    Continuous Liquid-Liquid Extraction With Hydrophobic Membrane

10.1.3.3.1     Follow the manufacturer's instructions for set-up.

10.1.3.3.2     Measure out each 1 L sample aliquot in a separate, clean
               graduated cylinder.  If the sample was received in a 1 L
               container, rinse the empty container with 50 mL of methylene
               chloride and add the rinsate to the continuous extractor.   If
               the sample was not received in a 1 L container, add  50 mL of
               methylene chloride to the continuous extractor.  Slowly
               transfer the aliquot to the continuous extractor. Adjust the pH
               to 2.0 with  1:1 H2SO4 and verify it with a pH meter or  narrow
               range pH paper.  Record the pH.

                                   D-28/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                                            Sample Preparation
10.1.3.3.3     Using a syringe or volumetric pipet, add 0.5 mL of the
               surrogate standard spiking solution (7.2.4.1) into the sample
               and mix well.

10.1.3.3.4     Rinse the graduated cylinder with 50 mL of methylene chloride
               and transfer the rinsate to the continuous extractor.

10.1.3.3.5     Add sufficient methylene chloride to the continuous extractor
               to ensure proper solvent cycling during operation.  Adjust the
               drip rate to 15 mL/minute (recommended); optimize the
               extraction drip rate.  Extract for a minimum of 6 hours. (NOTE:
               Due to the smaller volume of solvent used during the extraction
               process, some sample matrices (e.g., oily samples, samples
               containing a high concentration of surfactants) may create an
               emulsion which will consume the solvent volume, preventing the
               efficient extraction of the sample.  When this occurs, add
               additional solvent to assure efficient extraction of the
               sample, and extend the extraction time for a minimum of 6
               hours.  If the sample matrix prevents the free flow of solvent
               through the membrane, then the non-hydrophobic membrane
               continuous liquid-liquid type extractor must be used.)  Allow
               to cool, then detach the distillation flask.  Proceed to
               Section 10.2.

10.1.3.3.6     NOTE:  Some continuous liquid-liquid extractors are also capable
               of concentrating the extract within the extraction set-up.
               Follow the manufacturer's instructions for concentration when
               using this type of extractor.  Using the hydrophobic membrane
               type extractor, it may not be necessary to dry the extract with
               sodium sulfate.

10.1.3.4    NOTE: If low surrogate recoveries occur,  assure 1) the apparatus
            was properly assembled to prevent leaks;  2) the drip rate/solvent
            cycling was optimized; and 3) there was proper cooling for
            condensation of solvent.

10.1.3.5    NOTE: Alternate continuous liquid-liquid extractor types that meet
            the requirements of the SOW may also be used.  If using alternate
            extractors or design types, follow the manufacturer's instructions
            for set-up.

10.1.4   Soil/Sediment Samples

         Decant and discard any water layer on a sediment sample.  Mix samples
         thoroughly,  especially composited samples.  Discard any foreign
         objects such as sticks, leaves, and rocks.

10.1.4.1    pH Determination

            Transfer 50 g of soil/sediment to a 100 mL beaker.  Add 50 mL of
            water and stir for 1 hour.  Determine pH of sample with a pH meter
            while stirring.   Report pH value on appropriate data sheets.  If
            the pH of the soil/sediment is greater than 11 or less than 5,

                                  D-29/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Sample Preparation
            document this in the SDG Narrative.   Discard this portion of
            sample.  NOTE:   If limited sample volume is received use a smaller
            1:1 ratio of grams of soil/sediment  sample to mLs of water for the
            pH determination.   Note this in the  SDG Narrative.

10.1.4.2    Percent Moisture

            Immediately after weighing the sample for extraction, weigh 5-10 g
            of the soil/sediment into a tared crucible.  Determine the percent
            moisture by drying overnight at 105  °C.   Allow to cool in a
            desiccator before weighing.  Concentrations of individual analytes
            will be reported relative to the dry weight of soil/sediment.

               EQ. 4

          % Moisture = 9rams of wet sample - grams of dry sample x 10Q
                                 grains of wet sample
10.1.4.3    Mandatory Determination of Concentration Level

            The Contractor must determine whether a soil/sediment sample
            should be analyzed by the low or medium level soil/sediment
            method.  It is the responsibility of the Contractor to analyze the
            sample at the correct level.   Three approaches may be taken to
            determine whether the low level or medium 1-avel method must be
            fol. jwed.

            •   Assume the sample is low level and analyze a 30 g sample.

            •   Use the screening method in Appendix A to determine the
                appropriate method for analysis.

            •   Use an in-house laboratory screening procedure.  This
                procedure must be documented and available for review during
                on-site laboratory evaluation or when requested by the
                Technical Project Officer or Administrative Project Officer.

10.1.4.4    Low Level Soil/Sediment Samples

10.1.4,4.1     The following steps should be performed rapidly to avoid loss
               of the more volatile extractables.  Weigh approximately 30 g of
               sample to the nearest 0.1 g into a 400 mL beaker and add 60 g
               of anhydrous powdered or granulated sodium sulfate.  Mix well.
               The sample should have a sandy texture at this point.  Add 0.5
               mL of the surrogate standard spiking solution (Section 7.2.4.1)
               to the sample, then immediately add 100 mL of 1:1 methylene
               chloride-acetone.

10.1.4.4.2     Place the bottom surface of the tip of the 3/4 inch tapered
               disrupter horn about 1/2 inch below the surface of the solvent
               but above the sediment layer.
                                   D-30/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                                            Sample Preparation
10.1.4.4.3     Sonicate for 3 minutes using a 3/4 inch disrupter horn at full
               power (output control knob at 10) with pulse on and percent
               duty cycle knob set at 50.0 percent.  Do not use a microtip.
               NOTE:  These settings refer to the Model W-385.  When using a
               sonicator other than Model W-385, refer to the instructions
               provided by the manufacturer for appropriate output settings.
               Decant and filter extracts through Whatman #41 filter paper
               using vacuum filtration or centrifuge and decant extraction
               solvent.

10.1.4.4.4     Repeat the extraction two more times with 2 additional 100 mL
               portions of 1:1 methylene chloride-acetone.  Before each
               extraction, make certain that the sodium sulfate is free-
               flowing and not a consolidated mass.  As reg_uired, break up
               large lumps with a clean spatula, or, very carefully, with the
               tip of the unenergized probe.  Decant the extraction solvent
               after each sonication.  On the final sonication, pour the
               entire sample into the Buchner funnel and rinse with 1:1
               methylene chloride-acetone.

10.1.4.4.5  '   If the sample is to be screened following the low level
               preparation method prior to GPC,  refer to Appendix A,
               "Screening of Semivolatile Organic Extracts."

10.1.4.4.6     After screening, transfer the remainder of the 1 mL back to the
               total extract.  CAUTION: To minimize sample loss, autosamplers
               which pre-flush samples through the syringe should not be used.

10.1.4.5    Medium Level Soil/Sediment Samples

10.1.4.5.1     Transfer approximately 1 g (record weight to the nearest 0.1 g)
               of sample to a 20 mL vial.  Wipe the mouth of the vial with a
               tissue to remove any sample material.  Record the exact weight
               of sample taken.  Cap the vial before proceeding with the next
               sample to avoid any cross-contamination.

10.1.4.5.2     Add 2.0 g of anhydrous powdered or granulated sodium sulfate to
               the sample in the 20 mL vial, and mix well.

10.1.4.5.3     Surrogates are added to all samples, spikes, and blanks.  Add
               0.5 mL of surrogate spiking solution (Section 7.2.4.1) to the
               sample mixture.

10.1.4.5.4     Immediately add 9.5 mL of methylene chloride to the sample and
               disrupt the sample with the 1/8 inch tapered Microtip
               ultrasonic probe for 2 minutes at output control setting 5, in
               continuous mode (if using a sonicator other than Models W-375
               or W-385, contact the instrument manufacturer for appropriate
               output settings).   Before extraction, make certain that the
               sodium sulfate is free-flowing and not a consolidated mass.  As
               required, break up large lumps with a clean spatula, or, very
               carefully,  with the tip of the unenergized probe.
                                   D-31/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Sample Concentration
10.2  Concentrating the Extract

      NOTE: Low level soil/sediment samples prepared by the procedure
      described in Section 10.1.4.4 will result in extracts containing a
      mixture of acetone and methylene chloride.   Because all soil/sediment
      sample extracts MUST be subjected to GPC cleanup prior to analysis,  the
      majority of the  acetone must be removed from the extract,  otherwise it
      will have adverse effects on the GPC column.  To remove the acetone from
      the soil/sediment sample extract, follow the steps in Section 10.2.1
      then concentrate to 1 mL using the nitrogen evaporation technique in
      Section 10.2.2.2.

10.2.1   Concentration by K-D

10.2.1.1    Assemble a Kuderna-Danish (K-D) apparatus by attaching a 10 mL
            concentrator tube to a 500 mL evaporative flask.  Other
            concentration devices or techniques may be used in place of the K-
            D, if equivalency is demonstrated for all the semivolatile target
            compounds listed in Exhibit C.

10.2.1.2    For water samples, transfer the extract to a Kuderna-Danish (K-D)
            concentrator by pouring the extract through a drying column
            containing about 10 cm of anhydrous granular sodium sulfate.

10.2.1.2.1     For soil/sediment samples, directly transfer the extract to the
               K-D concentrator.

10.2.1.2.2     Rinse the Erlenmeyer flasks (for both water and soil/sediment
               samples) and the column (for water samples) with 20-30 mL of
               methylene chloride to complete the quantitative transfer.

10.2.1.3    Add one or two clean boiling chips to the evaporative flask and
            attach a three-ball Snyder column.   Pre-wet the Snyder column by
            adding about 1 mL methylene chloride to the top of the column.
            Place the K-D apparatus in a hot water bath (60 °C to 70 °C
            recommended) so that the concentrator tube is partially immersed
            in the hot water and the entire lower rounded surface of the flask
            is bathed with hot vapor.  Adjust the vertical position of the
            apparatus and the water temperature as required to complete the
            concentration in 10 to 15 minutes.  At the proper rate of
            distillation, the balls of the column will chatter actively, but
            the chambers will not flood with condensed solvent.  When the
            apparent volume of liquid reaches 1 mL, remove the K-D apparatus
            from the water bath and allow it to drain and cool for at Least  10
            minutes.  DO NOT ALLOW THE EVAPORATOR TO GO DRY.  Remove the
            Snyder column and rinse the flask and its lower joint into the
            concentrator tube with 1 or 2 mL of methylene chloride.  A 5 mL
            syringe is recommended for this operation.

10.2.1.4    For water samples which do not require GPC cleanup, proceed to
            final concentration of extract (Section 10.2.2).  Oily water
            samples require GPC.


                                   D-32/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                                          Sample Concentration
10.2.1.5    For water samples which require GPC, adjust the volume of the
            extract to 10.0 mL with methylene chloride and proceed with GPC
            cleanup (Section 10.3).

10.2.1.6    For soil/sediment samples, adjust the volume of the extract to
            10.0 mL with methylene chloride, and proceed with GPC cleanup
            (Section 10.3).

10.2.1.7    For water samples or soil/sediment samples which have undergone
            GPC, proceed to final concentration of extract (Section 10.2.2).

10.2.2   Final Concentration of Extract

         Two different concentration techniques are permitted to obtain the
         final extract volume:  Micro Snyder column and Nitrogen Evaporation
         Techniques.

10.2.2.1    Micro Snyder Column Technique

            Add another one or two clean boiling chips to the concentrator
            tube and attach a two-ball micro Snyder column.  Pre-wet the
            Snyder column by adding about 0.5 mL of methylene chloride to the
            top of the column.  Place the K-D apparatus on a hot water bath
            (60 °C to 70 °C recommended) so that the concentrator tube is
            partially immersed in the hot water.  Adjust the vertical position
            of the apparatus and the water temperature as required to complete
            the concentration in 5 to 10 minutes.   At the proper rate of
            distillation, the balls of the column will chatter actively, but
            the chambers will not flood with condensed solvent.  When the
            apparent volume of liquid reaches about 0.5 mL (0.4 mL for
            soil/sediment samples or water samples that have undergone GPC),
            remove the K-D apparatus from the water bath and allow it to drain
            for at least 10 minutes while cooling.  Remove the Snyder column
            and rinse the evaporative flask and its lower joint into the
            concentrator tube with 0.2 mL (0.1 mL for soil/sediment samples or
            water samples that have undergone GPC) of methylene chloride.
            Adjust the final volume to 1.0 mL (0.5 mL for soil/sediment
            samples and water samples that have undergone GPC) with methylene
            chloride.   Transfer the extract to the Teflon-sealed screw-cap
            bottle, label the bottle,  and store at less than 4 °C but not
            greater than 6 °C.

10.2.2.2    Nitrogen Evaporation Technique (Taken from ASTM Method D3086)

            The following method may be used for final concentration of the
            semivolatile extract instead of the procedure in Section 10.2.2.1.
            Place the concentrator tube in a warm water bath (30 °C to 35 °C
            recommended)  and evaporate the solvent volume to just below 1 mL
            (below 0.5 mL for soil/sediment samples and water samples that
            have undergone GPC)  by blowing a gentle stream of clean,  dry
            nitrogen (filtered through a column of activated carbon)  above the
            extract.  CAUTION:   Gas lines from the gas source to the
            evaporation apparatus must be stainless steel,  copper, or Teflon

                                  D-33/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Sample Cleanup (GPC)
            tubing.  New plastic tubing must not be used between the carbon
            trap and the sample, since it may introduce interferences.  The
            internal wall of the concentrator tube must be rinsed down several
            times with methylene chloride during the operation.  During
            evaporation the tube solvent level must be kept below the water
            level of the bath.  The extract must never be allowed to become
            dry.

10.2.2.3    Final Extract Volumes

10.2.2.3.1     Water

               For water samples that did not undergo GPC, the extract must be
               brought to a final volume of 1.0 mL with methylene chloride.
               For water samples that underwent GPC, the extract must be
               brought to a final volume of 0.5 mL with methylene chloride.
               (Concentrating the extract to 0.5 mL will result in no loss of
               sensitivity despite the volume of extract (5 mL) not recovered
               after GPC. )

10.2.2.3.2     Soil/Sediment

               Adjust the final volume to 0.5 mL with methylene chloride
               (concentrating the extract to 0.5 mL will result in no loss of
               sensitivity despite the volume of extract (5 mL) not recovered
               after GPC) .

10.2.2.3.3     Transfer the extract to a Teflon-sealed screw-cap bottle, label
               the bottle,  and store at less than 4 °C but not greater than 6
10.3  Sample Cleanup by GPC

10.3.1   Introduction

10.3.1.1    Gel Permeation Chromatography (GPC)  is a size exclusion cleanup
            procedure using organic solvents and hydrophobic gels in the
            separation of natural macromolecules .   The packing gel is porous
            and is characterized by the range or uniformity (exclusion range)
            of that pore size.   In the choice of gels, the exclusion range
            must be larger than the molecular size of the molecules to be
            separated.

10.3.1.2    GPC must be performed for all soil/sediment extracts.  GPC must be
            performed for water extracts that contain higher molecular weight
            contaminants that interfere with the analysis of the target
            analytes.

10.3.2   GPC Column Preparation

         The instructions listed below for GPC column preparation are for Bio
         Beads.  Alternative column packings may be used if 1) the column
         packings have equivalent or better performance than the Bio Beads and

                                   D-34/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatxles — Section 10
                                                                     Procedure
                                                          Sample Cleanup (GPC)
         meet the technical acceptance criteria for GPC calibration and GPC
         calibration check, 2) the column packings do not introduce
         contaminants/artifacts into the sample which interfere with the
         analysis of the semivolatile compounds.  Follow the manufacturer's
         instructions for preparation of the GPC column packing.

10.3.2.1    Weigh out 70 g of Bio Beads (SX-3).  Transfer them to a quart
            bottle with a Teflon-lined cap or a 500 mL separatory funnel with
            a large bore stopcock, and add approximately 300 mL of methylene
            chloride.  Swirl the container to ensure the wetting of all beads.
            Allow the beads to swell for a minimum of 2 hours.  Maintain
            enough solvent to cover the beads sufficiently at all times.  If a
            guard column is to be used, repeat the above with 5 g of Bio Beads
            in a 125 mL bottle or a beaker using 25 mL of the methylene
            chloride.

10.3.2.2    Turn the column upside down from its r.ormal position, and remove
            the inlet bed support plunger (the inlet plunger is longer than
            the outlet plunger).  Position and tighten the outlet bed support
            plunger as near the end as possible, but no closer than 5 cm
            (measured from the gel packing to the collar).

10.3.2.3    Raise the end of the outlet tube to keep the solvent in the GPC
            column, or close the column outlet stopcock.  Place a small amount
            of solvent in the column to minimize the formation of air bubbles
            at the base of poured column packing.

10.3.2.4    Swirl the bead/solvent slurry to get a homogeneous mixture and, if
            the wetting was done in a quart bottle, quickly transfer it to a
            500 mL separatory funnel with a large bore stopcock.  Drain the
            excess methylene chloride directly into the waste beaker, and then
            start draining the slurry into the column by placing the
            separatory funnel tip against the column wall.  This will help to
            minimize bubble formation.  Swirl occasionally to keep the slurry
            homogeneous.  Drain enough to fill the column.  Place the tubing
            from the column outlet into a waste beaker below the column, open
            the stopcock (if attached), and allow the excess solvent to drain.
            Raise the tube to stop the flow, and close the stopcock when the
            top of the gel begins to look dry.  Add additional methylene
            chloride to just rewet the gel.

10.3.2.5    Wipe any remaining beads and solvent from the inner walls of the
            top of the column with a laboratory tissue.  Loosen the seal
            slightly on the other plunger assembly (long plunger) and insert
            it into the column.   Make the seal just tight enough so that any
            beads on the glass surface will be pushed forward, but loose
            enough so that the plunger can be pushed forward.

            CAUTION:  Do not tighten the seal if beads are between the seal
            and the glass surface because this can damage the seal and cause
            leakage.
                                  D-35/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Sample Cleanup  (GPC)
10.3.2.6    Compress the column as much as possible  without applying
            excessive force.  Loosen the seal and gradually pull out the
            plunger.  Rinse and wipe off the plunger.  Slurry any remaining
            beads and transfer them into the column.  Repeat the step in
            Section 10.3.2.5 and reinsert the plunger.  If the plunger cannot
            be inserted and pushed in without allowing beads to escape around
            the seal, continue compression of the beads without tightening the
            seal, and loosen and remove the plunger as described.  Repeat this
            procedure until the plunger is inserted successfully.

10.3.2.7    Push the plunger until it meets the gel, then compress the column
            bed about four centimeters.

10.3.2.8    Pack the optional 5 cm column with approximately 5 g of pre-
            swelled beads (different guard columns may require different
            amounts).  Connect the guard column to the inlet of the analytical
            column.

10.3.2.9    Connect the column inlet to the solvent reservoir (reservoir
            should be placed higher than the top of the column) and place the
            column outlet tube in a waste container.  Placing a restrictor in
            the outlet tube will force air out of the column more quickly.  A
            restrictor can be made from a piece of capillary stainless steel
            tubing of 1/16" OD x 0.010" ID x 2".  Pump methylene chloride
            through the column at a rate of 5 mL/min for one hour.

10.3.2.10   After washing the column for at least one hour,  connect the column
            outlet tube without the restrictor,  to the inlet side of the UV
            detector.  Connect the system outlet to the outlet side of the UV
            detector.  A restrictor (same size as the one in Section 10.3.2.9)
            in the outlet tube from the UV detector will prevent bubble
            formation which causes a noisy UV baseline.

10.3.2.11   The restrictor will not affect flow rate.  After pumping methylene
            chloride through the column for an additional 1-2 hours, adjust
            the inlet bed support plunger until approximately 6-10 psi
            backpressure is achieved.   Push the plunger in to increase
            pressure or slowly pull outward to reduce pressure.

10.3.2.12   When the GPC column is not to be used for several days, connect
            the column outlet line to the column inlet to prevent column
            drying and/or channeling.   If channeling occurs, the gel must be
            removed from the column, re-swelled, and re-poured as described
            above.  If drying occurs,  methylene chloride should be pumped
            through the column until the observed column pressure is constant
            and the column appears wet.  Always recalibrate after column
            drying has occurred to verify retention volumes have not changed.
                                  D-36/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                                          Sample Cleanup (GPC)
10.3.3   Calibration of GPC

10.3.3.1    Summary of Calibration of GPC

            The GPC calibration procedure is based on monitoring the elution
            of standards with a UV detector connected to the GPC column.

10.3.3.2    Frequency of Calibration of GPC

            Each GPC system must be initially calibrated upon award of a
            contract, when the column is changed, when channeling occurs, and
            once every seven days when samples, including matrix spikes and
            matrix spike duplicates, and blanks are cleaned up using GPC.

10.3.3.3    Procedure for GPC Calibration

            The following instructions are for the Analytical Biochemical
            Laboratories system.  If you are using a different GPC system,
            consult your manufacturer's instruction manual for operating
            instructions.  A 2 mL injection loop may be used in place of a 5
            mL injection loop, in accordance with the manufacturer's
            instructions.

10.3.3.3.1     Verify the flow rate by collecting column eluate for 10 minutes
               in a graduated cylinder and measure the volume, which should be
               45-55 mL (4.5-5.5 mL/min).   Once the flow rate is within the
               range of 4.5-5.5 mL/min, record the column pressure (should be
               6-10 psi) and room temperature.  Changes in pressure,  solvent
               flow rate, and temperature conditions can affect analyte
               retention times and must be monitored.

10.3.3.3.2     Using a 10 mL syringe,  load sample loop #1 with calibration
               solution (Section 7.2.4.3).  With the ABC automated system, the
               5 mL sample loop requires a minimum of 8 mL of the calibration
               solution.  Use a firm continuous pressure to push the sample
               onto the loop.  Switch the valve so that GPC flow is through
               the UV flow-through cell.

10.3.3.3.3     Inject the calibration solution and obtain a UV trace showing a
               discrete peak for each component.  Adjust the detector and/or
               recorder sensitivity to produce a UV trace that meets the
               requirements in Section 10.3.4.  Differences between
               manufacturer's cell volumes and detector sensitivities may
               require a dilution of the calibration solution to achieve
               similar results.   An analytical flow-through detector cell will
               require a much less concentrated solution than the semi-prep
               cell and, therefore,  the analytical flow-through detector cell
               is not acceptable for use.

10.3.3.3.4     Using the information from the UV trace, establish appropriate
               collect and dump time periods to ensure collection of all
               target analytes.   Initiate column eluate collection just before
               elution of bis (2-ethylhexyl) phthalate and after the elution

                                  D-37/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Sample cleanup (GPC)
               of corn oil.  Stop eluate collection shortly after the elution
               of perylene.  Collection should be stopped before sulfur
               elutes.  Use a "wash" time of 10 minutes after the elution of
               sulfur.  Each laboratory is required to establish its specific
               time sequences.

10.3.3.3.5     Reinject the calibration solution after appropriate collect and
               dump cycles have been set, and the solvent flow and column
               pressure have been established.

10.3.3.3.6     Measure and record the volume of collected GPC eluate in a
               graduated cylinder.  The volume of GPC eluate collected for
               each sample extract processed may be used to indicate problems
               with the system during sample processing.

10.3.3.3.7     Analyze a GPC blank by loading 5 mL of methylene chloride into
               the GPC.  Concentrate the methylene chloride that passes
               through the system during the collect cycle using a Kuderna-
               Danish (K-D) evaporator.  Add internal standards at the
               appropriate concentration and analyze the concentrate by GC/MS.

10.3.4   Technical Acceptance Criteria for GPC Calibration

10.3.4.1    The GPC system must be calibrated at the frequency described in
            Section 10.3.3.2.  The UV trace must meet the following
            requirements:

            •   Peaks must be observed and should be symmetrical for all
                compounds in the calibration solution.

            •   Corn oil and the phthalate peaks must exhibit > 85.0 percent
                resolution.

            .•   The phthalate and methoxychlor peaks must exhibit > 85.0
                percent resolution.

            •   Methoxychlor and perylene peaks must exhibit > 85.0 percent
                resolution.

            •   Perylene and sulfur (if sulfur was added) peaks must not be
                saturated and must exhibit > 90.0 percent baseline resolution.

10.3.4.2    The solvent flow rate and column pressure must be within the
            ranges described in Section 10.3.3.3.1.

10.3.4.3    If the retention time shift is > 5.0 percent between calibrations,
            take corrective action.  Excessive retention time shifts are
            caused by the following:

            •   Poor laboratory temperature control or system leaks.

            •   An unstabilized column that requires pumping methylene
                chloride through it for several more hours or overnight.

                                   D-38/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                                         Extract Cleanup (GPC)


            •   Excessive laboratory temperatures causing outgassing of the
                methylene chloride.

10.3.4.4    A copy of the UV trace of the calibration solution must be
            submitted with the data for the associated samples.

10.3.4.5    The analyte concentrations in the GPC blank must contain less than
            the CRQL for all target compounds in Exhibit C (Semivolatiles),
            except phthalate esters, which must contain less than 5 times the
            CRQL.

10.3.5   Corrective Action for GPC Calibration

10.3.5.1    If the requirements in Section 10.3.4 cannot be met, the column
            may be cleaned by processing several 5 mL volumes of butylchloride
            through the system.  Butylchloride removes the discoloration and
            particles that may have precipitated out of the methylene chloride
            extracts.  If a guard column is being used, replace it with a new
            one.  This may correct the problem.  If column maintenance does
            not restore the performance of the column, the column must be
            repacked with new packing and recalibrated.  It may be necessary
            to obtain a new lot of Bio Beads if the column fails all criteria.

10.3.5.2    If the GPC blank exceeds the requirements in 10.3.4.5,  pump
            additional methylene chloride through the system for 1-2 hours.
            Analyze another GPC blank to ensure the system is sufficiently
            clean.  Repeat the methylene chloride pumping if necessary.

10.4  Sample Extract Cleanup by GPC

10.4.1   It is very important to have constant laboratory temperatures during
         an entire GPC run,  which could be 24 hours or more.   If temperatures
         are not constant, retention times will shift, and the dump and
         collect times determined by the calibration standard no longer will
         be appropriate.   The ideal laboratory temperature to prevent
         outgassing of the methylene chloride is 22 °C.

10.4.2   In order to prevent overloading of the GPC column, highly viscous
         sample extracts  must be diluted prior to cleanup.  Any sample extract
         with a viscosity greater than that of a 1:1 glycerol:water solution
         must be diluted  and loaded into several loops.  Similarly, extracts
         containing more  than 40 mg/mL of nonvolatile residue must be diluted
         and loaded into  several loops.  The nonvolatile residue may be
         determined by evaporating a 100 jiL aliquot of the extract to dryness
         in a tared aluminum weighing pan, or other suitable container.
         Systems using automated injection devices to load the sample on the
         column must be carefully monitored to assure that the required amount
         is being injected on the column.  Viscous extracts or extracts
         containing a large amount of non-volatile residue will cause problems
         with an automated injection system's ability to inject the proper
         amount of sample extract on a column.  After the sample extract has
         been processed,  the remaining sample extract in the injection vial
         must be checked  before proceeding with extract cleanup to assure that

                                   D-39/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 10
Procedure
Extract Cleanup (GPC)
         the proper amount was injected on the column.  If the proper amount
         of extract was not injected, the sample must be re-prepared at no
         additional cost to the Agency, and the sample extract must either be
         diluted and loaded into several loops or the sample extract must be
         injected manually.

         Note:  When multiple loops/runs are necessary for an individual
         sample, be sure to combine the all of the sample eluates collected
         from each run.

10.4.3   Frequency of GPC Sample Cleanup

         GPC cleanup must be performed once for each soil/sediment extract and
         for water extracts that contain high molecular weight contaminants
         that interfere with the analysis of the target analytes.  In
         addition, GPC must be performed for all associated blanks, matrix
         spikes and matrix spike duplicates.  If Cleanup procedure is
         inadequate, contact SMO.

10.4.4   Procedure for GPC Sample Cleanup

10.4.4.1    Particles greater than 5 microns may scratch the valve, which may
            result in a system leak and cross contamination of sample extracts
            in the sample loops.   To avoid such problems,  filter the extract
            through a 5 micron filter disc by attaching a syringe filter
            assembly containing the filter disc to a 10 mL syringe.  Draw the
            sample extract through the filter assembly and into the 10 /zL
            syringe.  Disconnect the filter assembly before transferring the
            sample extract into a small glass container, e.g., a 15 mL culture
            tube with a Teflon-lined screw cap.  Alternatively draw the
            extract into the syringe without the filter assembly.  Attach the
            filter assembly and force the extract through the filter and into
            the glass container.   Draw a minimum of 8 mL of extract into a 10
            mL syringe.

            Note:  Some GPC instrument manufacturer's recommend using a
            smaller micron size filter.  Follow the manufacturer's recommended
            operating instructions.

10.4.4.2    Introduction of particulates or glass wool into the GPC switching
            valves may require factory repair of the apparatus.

10.4.4.3    The following instructions are for the Analytical Biochemical
            Laboratories system.   If a different GPC system is being used,
            consult the manufacturer's instruction manual for operating
            instructions.  A 2 mL injection loop may be used in place of a
            5 mL injection loop.   If a 2 mL injection loop is used,
            concentrate the sample extract to 4 mL instead of 10 mL and then
            inject 2 mL instead of 5 mL.

10.4.4.4    Attach the syringe to the turn lock on the injection port.  Use
            firm, continuous pressure to push the sample onto the 5 mL sample
            loop.  If the sample is difficult to load, some part of the system

                                   D-40/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 10
                                                                     Procedure
                                  Final Concentration/Sample Analysis by GC/MS
            may be blocked.  Take appropriate corrective action.  If the back
            pressure  is normal  (6-10 psi) the blockage is probably in the
            valve.  Blockage may be flushed out of the valve by reversing the
            inlet and outlet tubes and pumping solvent through the tubes (this
            should be done before sample loading).

10.4.4.5    Approximately 2 mL of extract remains in the lines between the
            injection port and the sample loop; excess sample also passes
            through the sample loop to waste.  (ABC system)

10.4.4.6    After loading a loop, and before removing the syringe from the
            injection port, index the GPC to the next loop.  This will prevent
            loss of sample caused by unequal pressure in the loops.  After
            loading each sample loop, wash the loading port with methylene
            chloride  in a PTFE wash bottle to minimize cross-contamination.
            Inject approximately 10 mL of methylene chloride to rinse the
            common tubes.

10.4.4.7    After loading all the sample loops, index the GPC to 00 position,
            switch to the "RUN" mode and start the automated sequence.
            Process each sample using the collect and dump cycle time
            established in Section 10.3.3.3.3.

10.4.4.8    Collect each sample in a 250 mL Erlenmeyer flask, covered with
            aluminum  foil to reduce solvent evaporation,  or directly into a
            Kuderna-Danish (K-D) evaporator.  Monitor r. ample volumes
            col ected.  Changes in sample volumes collected may indicate one
            or more of the following problems:

            •   Change in solvent flow rate, caused by channeling in the
                column or changes in column pressure.

            •   Increase in column operating pressure due to the absorption of
                particles or gel fines onto either the guard column or the
                analytical column gel, if a guard column is not used.

            •   Leaks in the system or significant variances in room
                temperature.

10.5  Final Concentration

      Concentrate the extract as per Section 10.2.2.  Final volumes should be
      brought to the volumes stated in Section 10.2.2.3.

10.6  Sample Analysis by GC/MS

10.6.1   Sample extracts shall be analyzed only after the GC/MS system has met
         the instrument performance check,  initial calibration, and continuing
         calibration requirements.   The same instrument conditions must be
         employed for the analysis of samples as were used for calibration.

10.6.2   Internal standard solution is added to each sample extract.  Add 10
         /xL of internal standard solution to each accurately measured 1.0 mL

                                  D-41/SVOA                            OLM03.0

-------
 Exhibit  D  Semivolatiles  —  Section  10
 Procedure
 Sample Analysis  by  GC/MS
         of water  sample  extract.  For  soil/sediment samples  and water  samples
         subjected to GPC,  add  5 fiL of  internal standard solution to  each
         accurately measured 0.5 mL of  sample extract.  This  will result  in a
         concentration of 20 ng//*L of each internal standard.

         Note:  An alternate amount of  internal standard solution may be
         added, however, the internal standards must be added to maintain the
         required  20 ng//*L  of each internal standard in the sample extract.

10.6.3   Prior to  the addition  of internal standards, make any extract
         dilution  indicated by  characterization step.  Add internal standards
         after dilution.  Internal standards must be added to maintain  the
         required  20 ng/>L  of each internal standard in the extract volume.

10.6.4   Inject 2  /tL of the sample extract into the GC/MS.  This 2 /*L volume
         must contain 40 ng of  each internal standard.

10.6.5   Sample Dilutions

10.6.5.1    If the on-column concentration of any compound in any sample
            exceeds the initial calibration range, that sample extract  must be
            diluted,  the internal standard concentration must be readjusted,
            and the sample  extract must be reanalyzed.   Guidance in performing
            dilution and exceptions to this requirement are given below.

10.6.5.2    Use the results of the original analysis to determine the
            approximate dilution factor required to get the largest analyte
            peak within the initial calibration range.

10.6.5.3    The dilution factor chosen should keep the  response of the  largest
            peak for a target compound in the upper half of the calibration
            range of the instrument.

10.6.5.4    Do not submit data for more than two analyses,  i.e.,  the original
            sample extract and one dilution,  or if the  semivolatile screening
            procedure was employed,  from the most concentrated dilution
            analyzed and one further  dilution.
                                  D-42/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 11
                                                Data Analysis and Calculations
                                                    Qualitative Identification
11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Qualitative Identification

11.1.1   Identification of Target Compounds

11.1.1.1    The compounds listed in the Target Compound List (TCL), Exhibit C
            (Semivolatile), shall be identified by an analyst competent in the
            interpretation of mass spectra (see Exhibit A, Section 4.3.1) by
            comparison of the sample mass spectrum to the mass spectrum of the
            standard of the suspected compound.  Two criteria must be
            satisfied to verify the identifications.

            •   Elution of the sample analyte within GC relative retention
                time unit window established from the 12-hour calibration
                standard.

            •   Correspondence of the sample analyte and calibration standard
                component mass spectra.

11.1.1.2    For establishing correspondence of the GC relative retention time
            (RRT),  the sample component RRT must compare within ±0.06 RRT
            units of the RRT of the standard component.  For samples analyzed
            during the same 12-hour time period as the initial calibration
            standards, compare the analyte retention times to those from the
            50 ng calibration standard.  For reference, the standard must be
            run on the same shift as the sample.  If co-elution of interfering
            components prohibits accurate assignment of the sample component
            RRT from the total ion chromatogram, the RRT should be assigned by
            using extracted ion current profiles for ions unique to the
            component of interest.

11.1.1.3    For comparison of standard and sample component mass spectra, mass
            spectra obtained from a calibration standard on the Contractor's
            GC/MS meeting the daily instrument performance requirements for
            DFTPP are required.  Once obtained, these standard spectra may be
            used for identification purposes, only if the contractor's GC/MS
            meets the DFTPP daily instrument performance requirements.

11.1.1.4    The requirement for qualitative verification by comparison of mass
            spectra are as follows:

            •   All ions present in the standard mass spectrum at a relative
                intensity greater than 10.0 percent (most abundant ion in the
                spectrum equals 100.0 percent) must be present in the sample
                spectrum.

            •   The relative intensities of ions specified in the paragraph
                above must agree within ±20.0 percent between the standard
                and sample spectrum.   (Example: For an ion with an abundance
                of  50.0 percent in the standard spectrum, the corresponding
                sample ion abundance must be between 30.0 and 70.0 percent).
                                   D-43/SVOA                            OLM03.0

-------
 Exhibit D Semivolatiles — Section 11
 Data Analysis and Calculations
 Qualitative Identification
            •   Ions greater than 10.0 percent in the sample spectrum but not
                present in the standard spectrum must be considered and
                accounted for by the analyst making the comparison.  In Task
                II  (Exhibit A) the verification process should favor false
                positives.  All compounds meeting the identification criteria
                must be reported with their spectra.  When target compounds
                are below contract required guantitation limits (CRQL) but the
                spectrum meets the identification criteria, report the
                concentration with a "J".  For example, if the CRQL is 10 /tg/L
                and concentration of 3 /ig/L is calculated, report as "3J".

11.1.1.5    If a compound cannot be verified by all of the criteria in Section
            11.1.1  (11.1.1.1 through 11.1.1.4), but in the technical judgement
            of the mass spectra interpretation specialist the identification
            is correct, then the contractor shall report the identification
            and proceed with guantitation.

11.1.2   Identification of Non-Target Compounds

11.1.2.1    A library search shall be executed for non-target sample
            components for the purpose of tentative identification.  For this
            purpose, the NIST/EPA/NIH (May 1992 release cr later)  and/or Wiley
            (1991 release or later),  or equivalent mass spectral library shall
            be used.

11.1.2.2    Up to 30 organic compounds of greatest apparent concentration not
            listed in Exhibit C for the volatile and semivolatile fraction,
            excluding the surrogate and internal standard compounds,  shall be
            identified tentatively via a forward search of the NIST/EPA/NIH
            (May 1992 release or later)  and/or Wiley (1991 release or later),
            or equivalent mass spectral library.   The following are not to be
            reported; 1) Substances with responses less than 10 percent of the
            internal standard (as determined by inspection of the peak areas
            or heights); 2)  substances which elute earlier than 30 seconds
            before the first semivolatile compound listed in Exhibit C
            (Semivolatiles)  or three minutes after the last semivolatile
            compound listed in Exhibit C has eluted; and 3) volatile compounds
            listed in Exhibit C.   Only after visual comparison of sample
            spectra to spectra resulting from the library search(es)  will the
            mass spectral interpretation specialist assign a tentative
            identification.   NOTE:   Computer generated library search routines
            must not use normalization routines that would misrepresent the
            library or unknown spectra when compared to each other.

11.1.2.3    Peaks that are suspected to be straight-chain,  branched,  or cyclic
            alkanes, alone or part of an alkane series shall be library
            searched.  Documentation for the tentative identification must be
            supplied.  Alkanes are not counted as part of the 30 organic
            compounds described in section 11.1.2.2.

11.1.2.4    Peaks that are suspected to be aldol-condensation reaction
            products (i.e.,  4-methyl-4-hydroxy-2-pentanone and 4-methyl-3-
            pentene-2-one)  shall  be searched and reported and counted as part

                                  D-44/SVOA                            OLM03.1

-------
                                         Exhibit D Semivolatiles — Section 11
                                                Data Analysis and Calculations
                                                    Qualitative Identification
            of the 30 most intense non-target semivolatile compounds and are
            qualified with an "A" flag on Form I TIC.

11.1.2.5    Guidelines for Making Tentative Identifications

11.1.2.5.1     Major ions in the reference spectrum (ions greater than 10
               percent of the most abundant ion) should be present in the
               sample spectrum.

11.1.2.5.2     The relative intensities of the major ions should agree within
               ± 20 percent (Example:  For an ion with an abundance of 50
               percent in the standard spectrum, the corresponding sample ion
               abundance must be between 30 and 70 percent).

11.1.2.5.3     Molecular ions present in reference spectrum should be present
               in sample spectrum.

11.1.2.5.4     Ions present in the sample spectrum but not in the reference
               spectrum should be reviewed for possible background
               contamination or presence of co-eluting compounds.

11.1.2.5.5     Ions present in the reference spectrum but not in the sample
               spectrum should be reviewed for possible subtraction from the
               sample spectrum because of background contamination or co-
               eluting compounds.  NOTE:  Data system library reduction
               programs sometimes can create these disciepancies.

11.1.2.5.6     If,  in the technical judgement of the mass spectral
               interpretation specialist, no valid tentative  identification
               can be made, the compound should be reported as unknown.  The
               mass spectral interpretation specialist should give additional
               classification of the unknown compound, if possible (i.e.,
               unknown phthalate, unknown hydrocarbon, unknown acid type,
               unknown chlorinated compound).  If probable molecular weights
               can be distinguished, include them.

11.1.2.5.7     The Contractor shall report pesticide target compounds listed
               in Exhibit C (Pesticides) that appear as semivolatile
               tentatively identified compounds.

11.1.2.5.8     Straight-chain, branched, or cyclic alkanes are NOT to be
               reported as tentatively identifed compounds on FORM IF  (FORM I
               SV-TIC).  When the above alkanes are tentatively identifed, the
               concentration(s)  are to be estimated as described in Section
               11.2.2 and reported in the SPG narrative as alkanes, by class
               (i.e. straight-chain, branched, or cyclic; as  a series; as
               applicable).
                                  D-45/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 11
Data Analysis and Calculations
Calculations
11.2  Calculations

11.2.1   Target Compounds

11.2.1.1    Target compounds identified rhall be quantitated by the internal
            standard method.  The internal standard used shall be the one
            assigned to that analyte for guantitation (see Table 2).  The EICP
            area of primary characteristic ions of analytes listed in Tables 3
            and 4 are used for guantitation.

11.2.1.2    It is expected that situations will arise where the automated
            guantitation procedures in the GC/MS software provide
            inappropriate quantitations.  This normally occurs when there is
            compound co-elution, baseline noise, or matrix interferences.  In
            these circumstances, the Contractor must perform a manual
            guantitation.  Manual quantitations are performed by integrating
            the area of the quantitation ion of tne compound.  This
            integration shall only include the area attributable to the
            specific TCL compound.  The area integrated shall not include
            baseline background noise.  The area integrated shall not extend
            past the point where the sides of the peak intersect with the
            baseline noise.  Manual integration is not to be used solely to
            meet QC criteria, nor is it to be used as a substitute for
            corrective action on the chromatographic system.  Any instance of
            manual integration must be documented in the SDG Narrative.

11.2.1.3    In all instances where the data system report has been edited, or
            where manual integration or quantitation has been performed, the
            GC/MS operator must identify such edits or manual procedures by
            initialing and dating the changes made to the report, and shall
            include the integration scan range.  In addition, a hardcopy
            printout of the EICP of the quantitation ion displaying the manual
            integration shall be included in the raw data.  This applies to
            all compounds listed in Exhibit C  (Semivolatiles), internal
            standards and surrogates.

11.2.1.4    The requirements listed in 11.2.1.2 and 11.2.1.3 apply to all
            standards, samples  including MS/MSD, and blanks.

11.2.1.5    The relative response factor  (RRF) from the continuing calibration
            analysis is used to calculate the concentration  in the sample.
            For samples analyzed during the same 12-hour time period as the
            initial calibration standards, use the RRF  values from the  50 ng
            calibration standard.  Secondary ion quantitation is allowed ONLY
            when there are  sample interferences with the primary ion.   If
            secondary ion quantitation  is performed, document the reason in
            the SDG Narrative.  The area of a  secondary ion  cannot be used for
            the area of a primary ion unless a relative response factor  is
            calculated using the secondary  ion.

11.2.1.6    Calculate the concentration in  the sample using  the relative
            response factor (RRF) and the following equations.


                                   D-46/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles  — Section  11
                                                Data Analysis  and Calculations
                                                                   Calculations
11.2.1.6.1
Water
            EQ. 5
    Concentration
                                        (Ax) (I.) (Vt) (Df) (GPC)
                                      -                 (y. }
                Where,
                v,-   =
                RRF =
                GPC =
                Df  =
        Area of the characteristic ion for the compound to be measured
        Area of the characteristic ion for the internal standard
        Amount of internal standard injected in nanograms (ng)
        Volume of water extracted in milliliters (mL)
        Volume of extract injected in microliters (/iL)
        Volume of the concentrated extract in microliters (/^L)  (Vt =
        1,000 fj.L if sample was not subjected to GPC; Vt = 500 ^L if
        sample was subjected to GPC)
        Relative response factor determined from the 12-hour
        calibration standard
        GPC factor.
        GPC = 1.0 if water sample was not subjected to GPC;
        GPC = 2.0 if water sample was subjected to GPC.
        Dilution factor.   The dilution factor for analysis of water
        samples for semivolatiles by this method is defined as
        follows:
                          most cone, extract used to make dilution + uL clean solvent
                                 liL  most cone,  extract used to make dilution
                       If no dilution is performed, Df = 1.0
                                   D-47/SVOA
                                                        OLM03.0

-------
Exhibit D Semivolatiles — Section 11
Data Analysis and Calculations
Calculations
11.2.1.6.2     Soil/Sediment

               EQ. 6
               Where,

               Ax,  Is,  Ajs are as given for water,  above.
               Vt =  Volume of the concentrated extract in microliters (/iL)  (Vt =
                     500 fiL)
               Vj =  Volume of the extract injected in microliters (/uL)
               D  =  100 - % moisture
                          100
               Ws =  Weight of sample extracted in grams  (g)
               GPC=  GPC factor (GPC = 2.0 to account for GPC cleanup)
               RRF=  Relative response factor determined  from the 12-hour
                     calibration standard.
               Df =  Dilution factor. The dilution factor for analysis of
                     soil/sediment samples for semivolatiles by this method  is
                     defined as follows:

                     uL most cone, extract used to make dilution + ttL clean  solvent
                               /xL  most cone,  extract  used to make dilution

                     If no dilution is performed,  Df= 1.0

               The factor of 2.0 in the numerator is used to account for the
               amount of extract that is not recovered from the mandatory use
               of GPC cleanup.  Concentrating the extract collected after GPC
               to 0.5 mL maintains the sensitivity of the soil/ sediment
               method.

11.2.2   Non-Target Compounds

         An estimated concentration for non-target compounds tentatively
         identified shall be quantitated by the internal  standard method.  For
         guantitation, the nearest internal standard free of interferences
         shall be used.  The equations for calculating concentration are the
         same as equations 5 and 6.  Total area counts (or peak heights) from
         the total ion chromatograms are to be used for both the compounds to
         be measured and the internal standard.  A relative response factor
         (RRF) of one (1) is to be assumed.  The resulting concentration shall
         be qualified as "J" (estimated,  due to lack of a compound specific
         response factor), and "N" (presumptive evidence  of presence),
         indicating the quantitative and qualitative uncertainties associated
         with this non-target component.   An estimated concentration should  be
         calculated for all tentatively identified compounds as well as those
         identified as unknowns.
                                  D-48/SVOA                            OLM03.0

-------
                                         Exhibit D  Semivolatiles  —  Section 11
                                                Data  Analysis  and Calculations
                                                                   Calculations
11.2.3   CRQL Calculations

         If the CRQL  is  less than the CRQL  in Exhibit  C  (Semivolatiles),
         report the CRQL  listed  in Exhibit  C  (Semivolatiles) .

11.2.3.1    Water Samples

            EQ. 7

                     Adjusted   Contract   (Vx)
                              ~
                       CRQL   ~   CRQL       (V0) (Vc) (V
            Where ,
            Vt, Df, V0 and Vj  are  as  given  in  equation 6.
            Vx =  Contract sample volume (1,000 mL)
            Vy =  Contract injection volume (2 ^.L)
            Vc =  Contract concentrated extract volume  (1,000  jtL  if  GPC is  not
                  performed,  and  500 /iL if GPC was performed.
11.2.3.2    Soil/Sediment Samples

            EQ. 8
                     Adjusted __ Contract    (^ (VJ < V
                       CRQL       CRQL   x  (W } (V ) (V }
            Where,
            Vt,  Df,  Ws,  Vj and D are  as given  in  equation  7.
            Wx =  Contract sample weight  (30  g for low  level  and  1  g  for
                  medium level soil/sediment  samples)
            Vy =  Contract injection volume (2 /*L)
            Vc =  Contract concentrated extract  volume  (500 /tL, GPC is
                  required) .

11.2.4   Surrogate Recoveries

11.2.4.1    Calculate surrogate standard  recovery on all  samples, blanks,
            matrix spikes and matrix spike duplicates.  Determine if  recovery
            is within limits (see Table 7) and report on  appropriate  form.

11.2.4.2    Calculate the concentrations  of the  surrogate compounds using  the
            same equations as used for the target compounds.  Calculate the
            recovery of each surrogate using  the following equation:

            EQ.  10

                            Concentration (or amount) found
              ^Recovery  =
                            Concentration (or amount) spiked
                                   D-49/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 11
Data Analysis and Calculations
Technical Acceptance Criteria For Sample Analysis
11.3  Technical Acceptance Criteria for Sample Analysis

11.3.1   The samples must be analyzed on a GC/MS system meeting the DFTPP,
         initial calibration,  continuing calibration,  and blank technical
         acceptance criteria.   The sample must undergo cleanup procedures,
         when required, on a GPC meeting the technical acceptance criteria for
         GPC calibration.

11.3.2   The sample must be extracted and analyzed within the contract holding
         times.

11.3.3   The sample must have an associated method blank meeting the blank
         technical acceptance criteria.

11.3.4   The sample must have an acceptable surrogate  recovery.  The sample
         surrogate recovery will be considered unacceptable if the following
         occurs:

         •   Any surrogate has a recovery less than 10.0 percent.

         •   More than one base neutral surrogate or more than one acid
             surrogate is outside the surrogate acceptance windows listed in
             Table 7.

         NOTE:   These requirements do not apply to surrogates with advisory
         windows and to dilutions of samples-

11.3.5   The relative retention time of each surrogate must be within +0.06
         RRT units of its relative retention time in the continuing
         calibration standard.

11.3.6   The instrumental response (EICP area) for each of the internal
         standards must be within the inclusive range  of -50.0 percent and
         +100.0 percent of the response of the internal standards in the most
         recent continuing calibration analysis.

11.3.7   The retention time shift for each of the internal standards must be
         within ±0.50 minutes (30 seconds) between the sample and the most
         recent continuing calibration standard analysis.

11.3.8   Excluding those ions in the solvent front, no ion may saturate the
         detector.  No target compound concentration may exceed the upper
         limit of the 12-hour standard calibration range unless a more dilute
         aliquot of the sample extract is also analyzed according to the
         procedures in Section 10.6.5.

11.4  Corrective Action

11.4.1   Corrective Action for Sample Analysis

         The sample technical acceptance criteria must be met before data are
         reported.  Samples contaminated from laboratory sources, or sample
         results submitted not meeting the sample technical acceptance

                                   D-50/SVOA                           OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 11
                                                Data Analysis and Calculations
                                                             Corrective Action
         criteria, will require re-extraction and/or reanalysis at no
         additional cost to the Agency.

11.4.2   Corrective Action for Surrogate Recoveries Which Fail to Meet Their
         Acceptance Criteria  (Section 11.3.4, Table 7)

11.4.2.1    If the surrogate  compounds fail to meet their recovery acceptance
            criteria, check calculations, sample preparation logs, the
            surrogate compound spiking solutions, and the instrument
            operation.  If the calculations were incorrect, correct the
            calculations and  verify that the surrogate compound recoveries
            meet their acceptance criteria.  If the sample preparation logs
            indicate that the incorrect amount of surrogate compound spiking
            solution was added, then re-extract/reanalyze the sample after
            adding the correct amount of surrogate spiking solution.  If the
            surrogate compound spiking solution was improperly prepared,
            concentrated, or  degraded, re-prepare solutions and re-
            extract/reanalyze samples.  If the surrogate recoveries were
            outside the lower surrogate acceptance limit and the extract from
            the sample was cleaned up on a GPC using an automated injection
            system, the Contractor shall verify that the proper amount was
            injected on the GPC column.  If insufficient sample volume was
            injected on the GPC, the sample must be reprepared and reanalyzed
            at no additional cost to the Agency.  If the instrument
            malfunctioned,  correct the instrument problem and reanalyze the
            sample extract.  Verify that the surrogate recoveries meet their
            acceptance criteria.  If the instrument malfunction affected the
            calibrations, recalibrate the instrument before reanalyzing the
            sample extract.

11.4.2.2    If the above actions do not correct the problem, then the problem
            may be due to a sample matrix effect.  To determine if there was
            matrix effect,  take the following corrective action steps.

11.4.2.2.1     Re-extract and reanalyze the sample.  EXCEPTION: If surrogate
               compound recoveries in a sample used for a matrix spike and/or
               matrix spike duplicate were considered unacceptable, then it
               should be re-extracted/reanalyzed only if surrogate compound
               recoveries met the surrogate acceptance criteria in both the
               matrix spike and matrix spike duplicate analysis.

11.4.2.2.2     If the surrogate compound recoveries meet acceptance criteria
               in the re-extracted/reanalyzed sample, then the problem was
               within the Contractor's control.   Therefore, submit only data
               from the re-extraction/reanalysis.

11.4.2.2.3     If the surrogate compound recoveries fail to meet the
               acceptance criteria in the re-extracted/reanalyzed sample, then
               submit data from both analyses.   Distinguish between the
               initial analysis and the re-extraction/reanalysis on all
               deliverables,  using the suffixes in Exhibit B.
                                   D-51/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 11
Data Analysis and Calculations
Corrective Action
11.4.3   Corrective Action for Internal Standard Compound Responses Which Pail
         to Meet Their Acceptance Criteria (Section 11.3.5 and 11.3.6)

11.4.3.1    If the internal standards fail to meet their acceptance criteria,
            check calculations, the internal standard compound spiking
            solutions, and the instrument operation.   If the calculations were
            incorrect, correct the calculations and verify that the internal
            standard response met their acceptance criteria.  If the internal
            standard compound spiking solution was improperly prepared,
            concentrated, or degraded, re-prepare solutions and re-
            extract/reanalyze samples.  If the instrument malfunctioned,
            correct the instrument problem and reanalyze the sample extract.
            If the instrument malfunction affected the calibration,
            recalibrate the instrument before reanalyzing the sample extract.

11.4.3.2    If the above actions do not correct the problem, then the problem
            may be due to a sample matrix effect.  To determine if there was
            matrix effect, take the following corrective action steps.

11.4.3.2.1     Reanalyze the sample extract.   EXCEPTION:  If internal standard
               compound recoveries in a sample used for a matrix spike and/or
               matrix spike duplicate were outside t.he acceptance windows,
               then it should be reanalyzed only if internal standard compound
               recoveries met the surrogate acceptance criteria in both the
               matrix spike and matrix spike duplicate analysis.

11.4.3.2.2     If the internal standard compound recoveries meet acceptance
               criteria in the reanalyzed sample extract, then the problem was
               within the Contractor's control.  Therefore, submit only data
               from the reanalysis.

11.4.3.2.3     If the internal standard compound recoveries fail to meet their
               acceptance windows in the reanalyzed sample extract, then
               submit data from both analyses.  Distinguish between the
               initial analysis and the reanalysis on all deliverables, using
               the suffixes in Exhibit B.

11.4.4   Corrective Action for Surrogate Compounds Relative Retention
         Times/Internal Standard Compound Retention Times Outside Acceptance
         Criteria (Sections 11.3.5 and 11.3.6)

11.4.4.1    If the surrogate compounds relative retention times or internal
            standard compounds retention times are not within their acceptance
            criteria, check the instrument for malfunctions.  If the
            instrument malfunctioned, correct the instrument problem and
            reanalyze the sample extract. If the instrument malfunction
            affected the calibration, recalibrate the instrument before
            reanalyzing the sample extract.

11.4.4.2    If the above actions do not correct the problem, then the problem
            may be due to a sample "matrix effect".  To determine if there was
            matrix effect, take the following corrective action steps.
                                  D-52/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 11
                                                Data Analysis and Calculations
                                                             Corrective Action
11.4.4.2.1     Reanalyze the sample extract.   EXCEPTION:   If the surrogate
               compounds relative retention times or internal standard
               compounds retention times in a sample used for a matrix spike
               and/or matrix spike duplicate were outside the acceptance
               criteria, then it should be reanalyzed only if the surrogate
               compounds relative retention times and internal standard
               compounds retention times were within the acceptance criteria
               in both the matrix spike and matrix spike duplicate analysis.

11.4.4.2.2     If the surrogate compounds relative retention times and
               internal standard compounds retention times are within the
               acceptance criteria in the reanalyzed sample extract, then the
               problem was within the Contractor's control.  Therefore, submit
               only data from the reanalysis with the surrogate compounds
               relative retention times and the internal standard compound
               retention times within the acceptance limits.

11.4.4.2.3     If the surrogate compounds relative retention times or internal
               standard compounds retention times are outside the acceptance
               criteria in the reanalyzed sample extract, then submit data
               from both analyses.  Distinguish between the initial analysis
               and the reanalysis on all deliverables, using the suffixes in
               Exhibit B.

11.4.5   Corrective action for failure to meet instrument performance checks
         and initial and continuing calibration must be completed before the
         analysis of samples.
                                   D-53/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 12
Quality Control
Method Blanks
12.0  QUALITY CONTROL

12.1  Method Blanks

12.1.1   Summary of Method Blanks

         A method blank is a volume of a clean reference matrix (reagent water
         for water samples, or purified sodium sulfate for soil/sediment
         samples) that is carried through the entire analytical procedure.
         The volume or weight of the reference matrix must be approximately
         equal to the volume or weight of samples associated with the blank.
         The purpose of a method blank is to determine the levels of
         contamination associated with the processing and analysis of samples.

12.1.2   Frequency of Method Blanks

         Method blank extraction and analysis must be performed once for the
         following, whichever is most frequent, and analyzed on each GC/MS
         system used to analyze associated samples:

         •   Each SDG,  or

         •   Each 20 samples in an SDG,  including matrix spikes/matrix spike
             duplicates,  that are of a similar matrix (water, soil/sediment)
             or similar concentration (soil/sediment only),  or

         •   Whe  lever samples are extracted by tne same procedure (continuous
             liquid-liquid extraction or sonication).

12.1.3   Procedure for  Method Blank Preparation

         For semivolatile analyses, a method blank for water samples consists
         of 1 L volume  of reagent water spiked with 0.5 mL of the surrogate
         spiking solution.  For medium and low level soil/sediment samples, a
         method blank consists of 1 g and 30 g of sodium sulfate spiked with
         0.5 mL of the  surrogate spiking solution respectively.  Extract,
         concentrate, cleanup and analyze the blank according to procedures
         for water and  soil samples.

12.1.4   Technical Acceptance Criteria for Method Blank Analysis

12.1.4.1    All blanks  must be extracted and analyzed at the frequency
            described in  Section 12.1.2  on a GC/MS system meeting the DFTPP,
            initial calibration, and continuing calibration technical
            acceptance  criteria.

12.1.4.2    The blank must meet the sample acceptance criteria listed in
            Sections 11.3.4 through 11.3.7.

12.1.4.3    For all other target compounds,  the method blank must contain less
            than the CRQL of any single target compound (see Exhibit C).  A
            method blank  for semivolatile analysis must contain less than five
                                  D-54/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 12
                                                               Quality Control
                                                                 Method Blanks
            times  (5x) the contract required guantitation limit  (CRQL) of the
            phthalate esters listed in Exhibit C.

12.1.5   Corrective Action for Method Blanks

12.1.5.1    If a method blank does not meet the technical acceptance criteria
            for method blank analysis, the Contractor shall consider the
            analytical system to be out of control.

12.1.5.2    If contamination is the problem, then the source of the
            contamination must be investigated and appropriate corrective
            measures must be taken and documented before further sample
            analysis proceeds.  It is the Contractor's responsibility to
            ensure that method interferences caused by contaminants in
            solvent, reagents, glassware, and sample storage and processing
            hardware that lead to discrete artifacts and/or elevated baselines
            in the GC/MS be eliminated.  Samples associated with the
            contaminated blank must be re-extracted and reanalyzed at no
            additional cost to the Agency.

12.1.5.3    If surrogate recoveries in the method blank do not meet the
            acceptance criteria listed in 11.3.4 and Table 7,  first reanalyze
            the method blank.  If tho surrogate recoveries (except those with
            advisory limits)  do not meet the acceptance criteria after
            reanalysis, the method blank and all samples associated with that
            method blank must be re-extracted and reanalyzed at no additional
            cost to the Agency.

12.1.5.4    If the method blank does not meet internal standard response
            requirements listed in 11.3.6, check calculations, the internal
            standard spiking solutions, and the instrument operation.  If the
            calculations were incorrect, correct the calculations and verify
            that the internal standard responses meet their acceptance
            criteria.  If the internal standard compound spiking solution was
            improperly prepared,  concentrated,  or degraded,  re-prepare
            solutions and re-extract/reanalyze samples.   If the instrument
            malfunctioned, correct the instrument problem and reanalyze the
            method blank.  If the instrument malfunction affected the
            calibration,  recalibrate the instrument before reanalyzing the
            blank.  The Contractor shall resolve and document the resolution
            of the problem before proceeding with sample analysis.

12.1.5.5    If the method blank does not meet the retention time requirements
            for internal standards (11.3.7) or the surrogate (11.3.5), check
            the instrument for malfunction, and recalibrate.   Reanalyze the
            method blank.  Sample analyses cannot proceed until the method
            blank meets these requirements.
                                   D-55/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 12
Quality Control
MS/MSD


12.2  Matrix Spike/Matrix Spike Duplicate  (MS/MSD)

12.2.1   Summary of MS/MSD

         In order to evaluate the effects of the sample matrix on the methods
         used for semivolatile analyses, the Agency has prescribed a mixture
         of semivolatile target compounds to be spiked into two aliquots of a
         sample and analyzed in accordance with the appropriate method.

12.2.2   Frequency of MS/MSD Analyses

12.2.2.1    A matrix spike and matrix spike duplicate must be extracted and
            analyzed for every 20 field samples of a similar matrix in an SDG,
            whenever samples are extracted by the same procedure.

12.2.2.2    As part of the Agency's QA/QC program, water rinsate samples
            and/or field/trip blanks (field QC) may accompany soil/sediment
            samples and/or water samples that are delivered to the laboratory
            for analysis.   The Contractor shall not perform MS/MSD analysis on
            any of the field QC samples.

12.2.2.3    If the EPA Region designates a sample to be used as an MS/MSD,
            then that sample must be used.  If there is insufficient sample
            remaining to perform an MS/MSD, then the Contractor shall choose
            another sample on which to perform an MS/MSD analysis.  At the
            time the selection is made,  the Contractor shall notify the Region
            (through SMO)  that insufficient sample was received and identify
            the EPA sample selected for the MS/MSD analysis.  The  rationale
            for the choice of another sample other than the one designated by
            the Agency shall be documented in the SDG Narrative.

12.2.2.4    If there is insufficient sample remaining in any of the samples in
            an SDG to perform an MS/MSD, then the Contractor shall immediately
            contact SMO to inform them of the problem.  SMO will contact the
            Region for instructions.   The Region will either approve that no
            MS/MSD is required, or require that a reduced sample aliquot be
            used for the MS/MSD analysis.  SMO will notify the Contractor of
            the Region's decision.  The Contractor shall document  the decision
            in the SDG Narrative.

12.2.2.5    The Contractor will not be paid for MS/MSD analysis performed at
            greater frequency than is required by the contract.  If it appears
            that the Region has requested MS/MSD analysis at a greater
            frequency than required by the contract,  then the Contractor shall
            contact SMO.  SMO will contact the Region to determine which
            samples should have an MS/MSD performed on them.  SMO  will notify
            the Contractor of the Region's decision.   The Contractor shall
            document the decision in the SDG Narrative.

12.2.2.6    When a Contractor receives only a performance evaluation (PE)
            sample(s),  no MS/MSD shall be performed within that SDG.
                                  D-56/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Section 12
                                                               Quality Control
                                                                        MS/MSD
12.2.2.7    When a Contractor receives a performance evaluation (PE) sample as
            part of a larger SDG, a sample other than the PE sample must be
            chosen for the MS/MSD when the Region did not designate samples to
            be used for this purpose.  If the PE sample is an ampulated
            standard, the ampulated PE sample is not considered to be another
            matrix type.

12.2.3   Procedure for Preparing MS/MSD

12.2.3.1    Water Samples

            For water samples, prepare two additional 1 L aliquots of the
            sample chosen for spiking in two continuous extractors.  Add 0.5
            mL of surrogate spiking solution and 0.5 mL of the matrix spiking
            solution to each aliquot.  Extract,  concentrate, cleanup, and
            analyze the MS/MSD according to the procedures for water samples.

12.2.3.2    Soil/Sediment Samples - Low Level

            For low level soil/sediment samples, prepare two additional 30 g
            aliquots (record weight to nearest 0.1 g) of the sample chosen for
            spiking in the two 400 mL beakers.  Add 60 g of anhydrous powdered
            sodium sulfate to each aliqout.  Mix well.  Add 0.5 mL of the
            surrogate standard spiking solution and 0.5 mL of the matrix
            spiking solution to each aliquot, then immediately add 100 mL of
            1:1 methylene chloride-acetone.  Extract, concentrate, cleanup,
            and analyze the MS/MSD according to the procedures for low level
            soil samples.

12.2.3.3    Soil/Sediment Samples - Medium Level

            For medium level soil/sediment samples, prepare two additional 1 g
            aliquots (record weight to nearest 0.1 g) of the sample chosen for
            spiking in two 20 mL vials.  Add 2.0 g of anhydrous powdered
            sodium sulfate to each aliquot.  Mix well.  Add 0.5 mL of
            surrogate spiking solution, 0.5 mL of matrix spiking solution, and
            9.0 mL of methylene chloride.   Extract, concentrate, cleanup, and
            analyze the MS/MSD according to the procedures for medium level
            samples.

12.2.4   Dilution of MS/MSD

         Before any MS/MSD analysis, analyze the original sample,  then analyze
         the MS/MSD at the same concentration as the most concentrated extract
         for which the original sample results will be reported.  For example,
         if the original sample is to be reported at a 1:1 dilution and a 1:10
         dilution,  then analyze and report the MS/MSD at a 1:1 dilution only.
         However,  if the original sample is to be reported at a 1:10 dilution
         and a 1:100 dilution,  then the MS/MSD must be analyzed and reported
         at a 1:10 dilution only.  Do not  further dilute the MS/MSD samples to
         get either spiked or non-spiked analytes within calibration range.
                                  D-57/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 12
Quality Control
MS/MSD
         Dilution of the sample must be performed in accordance to the
         conditions in Section 10.6.5.

12.2.5   Calculations for MS/MSD

12.2.5.1    Calculate the recovery of each matrix spike compound in the matrix
            spike and matrix spike duplicate and report on the appropriate
            forms.  Calculate the concentrations of the matrix spike compounds
            using the same equations as used for target compounds (Equations 6
            and 7).  Calculate the recovery of each matrix spike compound as
            follows:

            EQ. 11
                     Matrix Spike Recovery = -	— x 100
            Where,

            SSR =   Spike Sample Result
            SR  =   Sample Result
            SA  =   Spike Added

12.2.5.2    Calculate the relative percent difference (RPD) of the recoveries
            of each compound in the matrix spike and matrix spike duplicate as
            follows:
            EQ.  12
                           RPD =   \MSR ~ MSDR\   x 100
                                   (MSR + MSDR}
            Where,

            RPD  =  Relative Percent Difference
            MSR  =  Matrix Spike Recovery
            MSDR =  Matrix Spike Duplicate Recovery

            The vertical bars in the formula above indicate the absolute value
            of the  difference,  hence RPD is always expressed as a positive
            value.
                                  D-58/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatxles — Section 12
                                                               Quality Control
                                                                        MS/MSD
12.2.6   Technical Acceptance Criteria for MS/MSD

12.2.6.1    All MS/MSD must be prepared and analyzed at a frequency described
            in Section 12.2.2.  All MS/MSD must be analyzed on a GC/MS system
            meeting DFTPP, initial and continuing calibration technical
            acceptance criteria and the method blank technical acceptance
            criteria.  The MS/MSD must undergo cleanup procedures, when
            required, on a GPC meeting the technical acceptance criteria for
            GPC calibration.

12.2.6.2    The MS/MSD must have an associated method blank meeting the blank
            technical acceptance criteria.

12.2.6.3    The MS/MSD must be extracted and analyzed within the contract
            holding time.

12.2.6.4    The retention time shift for each of  .he internal standards must
            be within ±0.50 minutes (30 seconds) between the MS/MSD sample and
            the most recent continuing calibration standard.

12.2.6.5    The relative retention time for the surrogate must be within
            ±0.06 RRT units of its standard retention time in the continuing
            calibration standard.

12.2.6.6    The limits for matrix spike compound recovery and RPD are given in
            Table 6.  As these limits are only advisory, no further action by
            the laboratory is required, however, frequent failures to meet the
            limits for recovery or RPD warrant investigation by the
            laboratory,  and may result in questions from the Agency.

12.2.7   Corrective Action for MS/MSD

         Any MS/MSD which fails to meet the technical acceptance criteria for
         MS/MSD must be reanalyzed at no additional cost to the Agency.
                                   D-59/SVOA                            OLM03.0

-------
Exhibix. D Semivolatiles — Sections 13 - 16
Method Performance/Pollution Prevention/Waste Management/References


13.0  METHOD PERFORMANCE

      Not Applicable.

14.0  POLLUTION PREVENTION

14.1  Pollution prevention encompasses any technique that reduces or
      eliminates the quantity or toxicity of waste at the point of generation.
      Numerous opportunities for pollution prevention exist in laboratory
      operation.  The EPA has established a preferred hierarchy of
      environmental management techniques that places pollution prevention as
      the management option of first choice.  Whenever feasible, laboratory
      personnel should use pollution prevention techniques to address their
      waste generation.  When wastes cannot be feasibly reduced at the source,
      the Agency recommends recycling as the next best option.

14.2  For information about pollution prevention that may be applicable to
      laboratories and research institutions, consult Less is Better:
      Laboratory Chemical Management, for Waste Reduction, available from the
      American Chemical Society's Department of Government Relations and
      Science Policy, 1155 16th Street N.W., Washington D.C.  20036, (202)
      872-4477.

15.0  WASTE MANAGEMENT

      The Environmental Protection Agency requires that laboratory waste
      management practices be conducted consistently with all applicable rules
      and regulations.    The Agency urges laboratories to protect the air,
      water, and land by minimizing and controlling all releases from hoods
      and bench operations,  complying with the letter and spirit of any sewer
      discharge permits and regulations,  and by complying with all solid and
      hazardous waste regulations, particularly the hazardous waste
      identification rules and land disposal restrictions.  Fcr further
      information on waste management consult The Waste Management Manual for
      Laboratory Personnel,  available from the American Chemical Society at
      the address listed in Section 14.

16.0  REFERENCES

      Not Applicable.
                                  D-60/SVOA                            OLM03.0

-------
                                          Exhibit D  Semivolatiles  —  Section  17
                                                     Tables/Diagrams/Flowcharts
 17.0  TABLES/DIAGRAMS/FLOWCHARTS

                                    Table 1

                   DFTPP Key Ions and Ion Abundance Criteria


               Mass   Ion Abundance Criteria

               51     30.0-80.0 percent of mass 198
               68     Less than 2.0 percent of mass 69
               69     Present
               70     Less than 2.0 percent of mass 69
               127    25.0 - 75.0 percent of mass 198
               197    Less than 1.0 percent of mass 198
               198    Base peak, 100 percent relative abundance
                      (see Note)
               199    5.0 - 9.0 percent of mass 198
               275    10.0 - 30.0 percent of mass 198
               365    Greater than 0.75 percent of mass 198
               441    Present but less than mass 443
               442    40.0 - 110.0 percent of mass 198
               443    15.0 - 24.0 percent of mass 442
Note:  All ion abundances MUST be normalized to m/z 198, the nominal base
peak, even though the ion abundance of m/z 442 may be up to 110 percent that
of m/z 198.
                                   D-61/SVOA                            OLM03.0

-------
.Exhibit D Semivolatiles — Section 17
Tables/Diagrams/Flowcharts










































CN

0)
rH
X)
10
EH





































































C
-H
TJ
c


no
Q)
«,
M




I |
•+-•


TO
y
rrt
Tl
y
(fl

C/3


nl
C

?n
jj
°p
i_i

0)
'"j

4-1
in
•o
• '
O
.^j


0)
CO













































c
o

4J
<0
4J
-H
4-1
C
(0
3
O

S-l
O
4-1

TJ
0)
C
en
-H
CO
CO
<

CO
0)
4J
10
D^
o
1-1
1-1
3
CO

TJ
C
(0

m
TJ
C
3
0
O4

O
o

4-1
01
1-1
(0
H

























V-
•o

0)
c
01

^1
}*l
0)
04



pj
•o"
0)
C
01
CO
>i
£
U
TJ
1
01
c
0)
1-1
x:
4-1
c
<0
C
0)
x;
a,



^_
TJ
1
01
c
0)
x;
4J
XI
o*
(0
C
0)
o
f£


CO
TJ
1
0)
c
0)
rH
«
x;
x:
a
m
2

T?
1
0)
C
0)
N
c
0)
X!
0
^
0
rH
("I
U
-H
Q
1

•*
rH



1
rH
^
4J
O
O
1
c
1
-H
Q







01
C
0)
S-l
a,
i
CM
1
O
to
4J
-H
C
-H
Q
1
<£>
•*
*3*


|
0
rH
O
^
U
o
to
O
rH
XI
o
(0
X
0)
hC




0)
c
0)
M
c
0)
X!
S-l
4-1
-H
2















rH
0
C
Q)
x;
Q4
1 1 0)
to to C l l i
O 0 01 ro 0) *~* •^•* 0)
0)3 3 S4 » c -H x: c

(0 MH <4H CLi - S-i X! C (0 O
rH «— - *—* «—* rH ^1 •» Q) ^" (0
10 Xt Q) *^ 0) 10 '*-' Pi D> rH O S-1
XI ^-* C "™^ C ^-^ O 1 ^~" ^i N XI
-POO)OO)OC'-OS-lC4-i
XJNX!NXiHQ)TJNO)Q)C
O4 C4-lC4JCTJOCQ4Xt'0
o> c  c ai -H

0)
1 4->
O (0 .j.
1 to 0) 1 rH £
rHCUOO) CrHlO TJ
>i 4-> r-i c o) >. x: i
C rH O TJ *•••" (0 4-) jC 0) ^i *""*
fl) (0 *H -H (0 S-l 0) PJ C C S4
xixia N — ' x: i — oioiJ-i
•H 4J 1 CO4-)CNrHCQX«3
>i xi * 0)NC^>~>i>iO4(n
3 - 0) -H 0) XI 0)
m ro m xi x; o EH
^H
0 1 0) rH to
C -H C >i 0) 1 0)
0) TJ -H C XI 1 O C 1
x: 1 E 0) 4-> 0 S-i 0)0)rH
O4O>0X!0)S-40)0 OJS^C^i
rH BlrH Cl4 rH 0 CrHrHrHXI 0)4J
I^i O ^i O 0 t~~^ 0) ,^I 0 0 -J-J O 13
c| Vw C £ C .C N O C t*] C cO f^
-P^J(UOQJOCW Q) 4-J 1
i i ai oi (0 x: c -H
j (Q Q


i i
0 O
to to 0) 0) 01
0 O 1 C C C
0) rH rH £ -H HI -r-l (1)
CXI XI O4 rH OJ-HrHC
0)O O <0 -H 4J >i -H 0)
•H -H -H c 0) c i to x; c x:
TJS-|rHS^rHOC(0rHrH4J(04J
(0EHO6HOS-IQJO ^fOXIOXI
4-11 Cl COrHS-jX!X!O4S^O4
CvD Q)L^1 0)rH C04J4J4J C04J 10
Q) ^XI ^XIXtX!-HO)XIC-HC
O4^ 04^1* O4O4-1Z £ OwCDZS 0)
1 1 -H O 1 O
CN CN CN CNQ i S-lO)l-4 XIQJO)(0t
oioix; oeo uccco
C XI 4-1 rH -^^ rH -H 0) 0) (0 S-l
O Q40)rH,C ^I^MI — 1 S-l N rH O O
S^O£OOXUOEHCt0S-irH
OSJ-HCI o--tci oixiox;
CU -H I jC *"• ' 4-1 1 f* ^ O XI XI (0
O2S^* O4CO 0)^* CUCN S-i Pi c_) x
CO 1 - -H - <. 10 1 0)
MCNCN XI CN rH Z^TK


0) 0) 0)
— c c c
rH 0) 0) 0)
>! N N M — 0)
XI C C C 10) 1 C
Jj rHOIOlOlrHtHCrHC 10
0) oxixixio^<0oi oix:
O COOOCI&C -H C 4-1
S4 0)S^S4S-|Q)Q100)Q 'H Q)
O XIOOOXI-HS-iXIISO
rH CU-HrHrHaXlOjSllOflSj
XI OXIXIXIrH^OrHBlrHO
OS-iS-jUUU^iXS^^iO^irH
1 OJO-H-H-HXIOOXIS^QjXI
CNXIrHQQQ4J 1 rH4->4-lOO
— '4JXII 1 1Q)*XIO)-^S-|(0
CO (UUf^^tCNSCNOSZ Q
o
S-l

-H
c
-H
Q
1

h.
CN







0)
c
0)
rH
(Tl
(U
x;
4->















^^
S-l
S-l
3
en
•«„*










































0)
c
0) 0) rH
3 4J 1 >t
rH (0 rH Q) C
O rH >1 CO)
jj (0 c .H x;
0 x; 0) sj rH &,
to 4-1 XI 0) -H -H
4J x: 04 x: c xi
•H Qj O 4-> 0) (0 O
C rH to 0) C O 1-1
•H>,OrHO)SjO-~
Q.CrH >,S-I4-) 3 1-4
i 4-> x: c o ---I rH to

- -H 1 x: rH I I [11
CN Q ^" P[ p* t *3* CN *^^

in
TJ
1
0)
c
0)
N
c
01
XI —
O *-J
S-l S-l
4-1 3
-H CO
z ~

T/
1
0)
c
0)
N
c
0)
X!
0
to
O

XI !H
0 to
•H 3
a co
i -~-
CN
^
rH







































































Tf
c
3
o
a


^
fll
w
"Irt
(-JN
W
o
ii


0)

II

ti
j^

CO
                                  D-62/SVOA
OLM03.0

-------
                               Exhibit D Semivolatiles  —  Section  17
                                          Tables/Diagrams/Flowcharts
                          Table 3

        Characteristic Ions for Internal Standards
                 for  Semivolatile  Compounds
                              Primary
                           Quantitation
INTERNAL STANDARDS	Ion	Secondary Ions

l,4-Dichlorobenzene-d4          152         115
Naphthalene-d8                  136         68
Acenaphthene-dlO                164         162, 160
Phenanthrene-dlO                188         94, 80
Chrysene-dl2                    240         120, 236
Perylene-dl2                    264	260, 265	
                         D-63/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Section 17
Tables/Diagrams/Flowcharts
                                   Table 4

                    Characteristic Ions for Semivolatile
                       Target Compounds and Surrogates
Parameter
Phenol
bis ( 2-Chloroethyl ) ether
2-Chlorophenol
1, 3-Dichlorobenzene
1 , 4-Dichlorobenzene
1 , 2-Dichlorobenzene
2-Methylphenol
2 , 2 '-oxybis ( 1-Chloropropane)
4-Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4-Dimethylphenol
bis ( 2-Chloroethoxy ) methane
2 , 4-Dichlorophenol
1,2, 4-Trichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-methylphenol
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2,4, 5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
Dimethyl phthalate
Acenaphthylene
3 -Nit roan i line
Acenaphthene
2 , 4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 , 4-Dinitrotoluene
2 , 6-Dinitrotoluene
Diethylphthalate
4-Chlorophenyl-phenylether
Fluorene
4-Nitroaniline
4, 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenylether
Hexachlorobenzene
Pentachlorophenol
Primary
Quant itat ion
Ion
94
93
128
146
146
146
108
45
108
70
117
77
82
139
107
93
162
180
128
127
225
107
142
237
196
196
162
65
163
152
138
153
184
109
168
165
165
149
204
166
138
198
169
248
284
266
Secondary Ion(s)
65, 66
63, 95
64, 130
148, 113
148, 113
148, 113
107
77, 79
107
42, 101, 130
201, 199
123, 65
95, 138
65, 109
121, 122
95, 123
164, 98
182, 145
129, 127
129
223, 227
144, 142
141
235, 272
198, 200
198, 200
164, 127
92, 138
194, 164
151, 153
108, 92
152, 154
63, 154
139, 65
139
63, 182
89, 121
177, 150
206, 141
165, 167
92, 108
182, 77
168, 167
250, 141
142, 249
264, 268
                                  D-64/SVOA
OLM03.0

-------
                                   Exhibit  D  Semivolatiles  —  Section 17
                                              Tables/Diagrams/Flowcharts
                        Table 4 (cont'd.)

              Characteristic Ions for Semivolatile
                 Target Compounds and Surrogates
Parameter
                                Primary
                                Quant itat ion
                                Ion
Secondary Ion(s)
Phenanthrene                    178
Anthracene                      178
Carbazole                       167
Di-n-butylphthalate             149
Fluoranthene                    202
Pyrene                          202
Butylbenzylphthalate            149
3,3'-Dichlorobenzidine          252
Benzo(a)anthracene              228
bis(2-Ethylhexyl)phthalate      149
Chrysene                        228
Di-n-Octyl phthalate            149
Benzo(b)fluoranthene            252
Benzo(k)fluoranthene            252
Benzo(a)pyrene                  252
Indeno(l,2,3-cd)pyrene          276
Dibenzo(a,h)anthracene          278
Benzo(g,h,i)perylene            276
SURROGATES
Pnenol-d"                       99
2-Fluorophenol                  112
2,4,6-Tribromophenol            330
Nitrobenzene-d5                 82
2-Fluorobiphenyl                172
Terphenyl-dl4                   244
2-Chlorophenol-d4               132
l,2-Dichlorobenzene-d4          152
179, 176
179, 176
166, 139
150, 104
101, 100
101, 100
91, 206
254, 126
229, 226
167, 279
226, 229

253, 125
253, 125
253, 125
138, 227
139, 279
138, 277

42, 71
64
332, 141
128, 54
171
122, 212
68, 134
115, 150
                            D-65/SVOA
                 OLM03.0

-------
Exhibit D Semivolatiles — Section 17
Tables/Diagrams/Flowcharts
                                   Table  5

         Relative Response Factor Criteria for Initial and Continuing
         Calibration of  Semivolatile Target Compounds and  Surrogates
Semivolatile
Compounds
Phenol
bis(2-Chloroethyl ) ether
2-Chlorophenol
1 , 3-Dichlorobenzene
1 , 4-Dichlorobenzene
1 , 2-Dichlorobenzene
2-Methylphenol
2,2' -oxybis ( 1-Chloropropane )
4-Methylphenol
N-Nitroso-di-n-propylamine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2 , 4-Dimethylphenol
bis ( 2-Chloroethoxy)methane
2 , 4-Dichlorophenol
1,2, 4-Trichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-methylphenol
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4, 6-Trichlorophenol
2,4, 5-Trichlorophenol
2 -Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
Acenaphthylene
3 -Nit roan i line
2 , 6-Dinitrotoluene
Acenaphtuene
2 , 4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2 , 4-Dinitrotoluene
Diethylphthalate
4-Chlorophenyl-phenylether
Fluorene
4-Nitroaniline
4 , 6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenylether
Hexachlorobenzene
Pentachlorophenol
Phenanthrene
llinimum
RRF
0.800
0.700
0.800
0.600
0.500
0.400
0.700
0.010
0.600
0.500
0.300
0.200
0.400
0.100
0.200
0.300
0.200
0.200
0.700
0.010
0.010
0.200
0.400
0.010
0.200
0.200
0.800
0.010
0.010
0.900
0.010
0.200
0.900
0.010
0.010
0.800
0.200
0.010
0.400
0.900
0.010
0.010
0.010
0.100
0.100
0.050
0.700
Maximum
%RSD
20.5
20.5
20.5
20.5
20.5
20.5
20.5
none
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
20.5
none
none
20.5
20.5
none
20.5
20.5
20.5
none
none
20.5
none
20.5
20.5
none
none
20.5
20.5
none
20.5
20.5
none
none
none
20.5
20.5
20.5
20.5
Maximum
%Diff
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0
none
none
±25.0
±25.0
none
±25.0
±25.0
±25.0
none
none
±25.0
none
±25.0
±25.0
none
none
±25.0
±25.0
none
±25.0
±25.0
none
none
none
±25.0
±25.0
±25.0
±25.0
                                  D-66/SVOA
OLM03.0

-------
                                Exhibit D Semivolatiles — Section 17
                                           Tables/Diagrams/Flowcharts
                      Table  5  (cont'd.)

Relative Response Factor Criteria for Initial and Continuing
 Calibration  of  Semivolatile Target Compounds and Surrogates
Semivolatile
Compounds
Anthracene
Carbazole
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3' -Dichlorobenzidine
Benzo ( a ) anthracene
bis ( 2-Ethylhexyl ) phthalate
Chrysene
Di-n-octylphthalate
Benzo (b) f luoranthene
Benzo(k) f luoranthene
Benzo ( a ) pyrene
Indeno{ 1,2, 3-cd) pyrene
Dibenzo(a, h) anthracene
Benzo (g,h, i)perylene
SURROGATES
Nitrobenzene-d5
2-Fluorobiphenyl
Terphenyl-dl4
Phenol-d5
2-Fluorophenol
2,4, 6-Tribromophenol
2-Chlorophenol-d4
1 , 2-Dichlorobenzene-d4
Minimum
RRF
0.700
0.010
0.010
0.600
0.600
0.010
0.010
0.800
0.010
0.700
0.010
0.700
0.700
0.700
0.500
0.400
0.500

0.200
0.700
0.500
0.800
0.600
0.010
0.800
0.400
Maximum
%RSD
20.5
none
none
20.5
20.5
none
none
20.5
none
20.5
none
20.5
20.5
20.5
20.5
20.5
20.5

20.5
20.5
20.5
20.5
20.5
none
20.5
20.5
Maximum
%Diff
±25.0
none
none
±25.0
±25.0
none
none
±25.0
none
±25.0
none
±25.0
±25.0
±25.0
±25.0
±25.0
±25.0

±25.0
±25.0
±25.0
±25.0
±25.0
none
±25.0
±25.0
                         D-67/SVOA
OLM03.0

-------
Exhibit D Semivolatiles — Section 17
Tables/Diagrams/Flowcharts
                                   Table 6

                          Matrix Spike Recovery and
                      Relative Percent Difference  Limits
 Compound
% Recovery
  Water
RPD
Water
% Recovery
  Soil/
Sediment
RPD
Soil/
Sediment
 Phenol                           12-110      42
 2-Chlorophenol                   27-123      40
 1,4-Dichlorobenzene              36-97       28
 N-Nitroso-di-n-propylamine       41-116      38
 1,2,4-Trichlorobenzene           39-98       28
 4-Chloro-3-methylphenol          23-97       42
 Acenaphthene                     46-118      31
 4-Nitrophenol                    10-80       50
 2,4-Dinitrotoluene               24-96       38
 Pentachlorophenol                9-103       50
 Pyrene                           26-127      31
                      26-90
                      25-102
                      28-104
                      41-126
                      38-107
                      26-103
                      31-137
                      11-114
                      28-89
                      17-109
                      35-142
                      35
                      50
                      27
                      38
                      23
                      33
                      19
                      50
                      47
                      47
                      36
                                  D-68/SVOA
                                    OLM03.0

-------
                                        Exhibit D Semivolatiles — Section 17
                                                   Tables/Diagrams/Flowcharts
                                   Table  7

                          Surrogate Recovery  Limits
                                   %Recovery            %Recovery
COMPOUND                           Water                Soil/Sediment

Nitrobenzene-d5 (Base/Neutral)     35-114               23-120
2-Fluorobiphenyl (Base/Neutral)    43-116               30-115
Terphenyl-dl4 (Base/Neutral)       33-141               18-137
Phenol-d5 (Acid)                   10-110               24-113
2-Fluorophenol (Acid)              21-110               25-121
2,4,6-Tribromophenol (Acid)        10-123               19-122
2-Chlorophenol-d4 (Acid)           33-110 (advisory)    20-130 (advisory)
l,2-Dichlorobenzene-d4             16-110 (advisory)    20-130 (advisory)
(Base/Neutral)
                                 D-69/SVOA                            OLM03.0

-------
Exhibit  D  Semivolatiles  — Appendix A
Screening  of  Semivolatile Organic Extracts
APPENDIX A -  SCREENING OF SEMIVOLATILE ORGANIC EXTRACTS

1.0   SCOPE AND APPLICATION

1.1   The analytical method described in this section is designed to
      characterize soil/sediment samples from hazardous waste sites for the
      concentration level of organics.

1.2   The method involves the preparation of soil/sediment samples which may
      contain organic compounds at a level greater than 10,000 jtg/kg and
      screening with a gas chromatograph equipped with a flame ionization
      detector (GC/FID).

1.3   For soil/sediment samples, the results of the screen are used to
      determine which of the two sample preparation procedures (low or medium)
      is required, and to determine appropriate dilution factor for GC/MS
      analysis.  The results of the screen may also be used to assist the
      analyst in performing Gel Permeation Chromatography (GPC) cleanup
      procedures on extracts of either water or soil/sediment samples.

1.4   The procedure is designed to allow a g_uantitation limit for screening
      purposes as low as 10,000 /ug/kg for extractable organics.  For analysis
      purposes, the quantitation limit is 10,000 ^g/kg for extractable
      organics.  Some samples may contain high concentrations of compounds
      that interfere with the analysis of other components at lower levels;
      the guantitation limits in those cases may be significantly higher.

1.5   These ext iction and preparation procedures were developed for rapid and
      safe handling of high concentration hazardous waste samples.  The design
      of the methods thus does not stress efficient recoveries or low limits
      of quantitation of all components.  Rather, the procedures were designed
      to screen,  at moderate recovery and sufficient sensitivity, a broad
      spectrum of organic chemicals.   The results of the analyses thus may
      reflect only a minimum of the amount actually present in some samples.

2.0   SUMMARY OF METHOD

2.1   It is mandatory that all soil/sediment samples be characterized as to
      concentration level so that the appropriate analytical protocol is
      chosen to ensure proper quantitation limits for the sample.  Note that
      the terms "low level" and "medium level" are descriptions of the
      concentration ranges that are encompassed by the "low" and "medium"
      level procedures.

2.2   The laboratory is at liberty to select the method of characterization.
      The following two screening methods may  be used for soil/sediment
      sample characterization:

      •   Screen an aliquot from the "low level" 30 g extract or an aliquot
          from the "medium level" 1 g extract.

      •   Screen using GC/FID as the screening instrument.
                                   D-70/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Appendix A
                                    Screening of Semivolatile Organic Extracts
2.3   The concentration ranges covered by these two procedures may be
      considered to be approximately 330 /*g/kg - 10,000 /*g/kg for the low
      level analysis and >10,000 /xg/kg for medium level analysis for
      semivolatile extractables.

2.4   Sample Preparation

2.4.1    Low Level Soil/Sediment

         A 30 g portion of soil/sediment is mixed with anhydrous powdered
         sodium sulfate and extracted with 1:1 methylene chloride/acetone
         using an ultrasonic probe.
         5.0 mL of the 300 mL (approximate) total extract is concentrated to
         1.0 mL and screened.  If the original sample concentration is >10,000
         /xg/Kg, the 30 g extract is discarded and the medium level preparation
         procedure followed.

2.4.2    Medium Level Soil/Sediment

         Approximately 1 g portions of soil/sediment are transferred to vials,
         mixed with anhydrous powdered sodium sulfate and extracted with
         methylene chloride.  5.0 mL of the 10.0 mL extract is concentrated to
         1.0 mL and screened.  If the sample concentration is <10,000 ^tg/kg
         the 1 g extract is discarded and the low level method is followed.

2.5   GC/FID Screening

      The concentrated extracts of soil/sediment or water samples are screened
      on a gas chromatograph/flame ionization detector using a fused silica
      capillary column (FSCC) for semivolatile priority pollutants and related
      organics.  The results of these screens will determine whether
      sufficient quantities of pollutants are present to warrant analysis by
      the medium level protocol.

3.0   INTERFERENCES

      Method interferences may oe caused by contaminants in solvents,
      reagents, glassware, and other sample processing hardware that lead to
      discrete artifacts and/or elevated baselines in the chromatograms.  All
      of these materials routinely must be demonstrated to be frae from
      interferences under the conditions of the analysis by running laboratory
      method blanks.  Matrix interferences may be caused by contaminants that
      are co-extracted from the sample.  The extent of matrix interferences
      will vary considerably from source to source.

4.0   SAFETY

4.1   The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely determined; however, each chemical should be treated
      as a potential health hazard.  Exposure to these reagents should be
      reduced to the lowest possible level.  The laboratory is responsible for
      maintaining a current awareness file of OSHA regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
      file of data handling sheets should be made available to all personnel

                                   D-71/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Appendix A
Screening of Semivolatile Organic Extracts


      involved in these analyses.  Specifically, concentrated sulfuric acid
      presents some hazards and is moderately toxic and extremely irritating
      to the skin and mucous membranes.  Use these reagents in fume hoods
      whenever possible, and if eye or skin contact occurs, flush with large
      volumes of water.

4.2   Always wear safety glasses or a shield for eye protection, protective
      clothing and observe proper mixing when working with these reagents.

5.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, catalog and part numbers are for illustrative
      purposes only.  No endorsement is implied.  Equivalent performance may
      be achieved using equipment and supplies other than those specified
      here, but demonstration of equivalent performance meeting the
      requirements of this SOW is the responsibility of the contractor.

5.1   Glassware

5.1.1    Continuous Liquid-Liquid Extractors - equipped with Teflon or glass
         connecting joints and stopcocks requiring no lubrications (Hershberg-
         Wolf Extractor, Ace Glass Company, Vineland, NJ.  P/N 6841 10 or
         equivalent).

5.1.2    Beakers - 400 mL

5.1.3    Syringes - 0.5 mL

5.1.4    Glass Scintillation Vials - at least 20 mL, with screw cap and Teflon
         or aluminum foil liner.

5.1.5    Vials and Caps - 2 mL capacity for GC auto sampler.

5.1.6    Disposable Pipets - Pasteur,  1 mL

5.1.7    Drying Column - 19 mm ID chromatographic column with coarse frit
         (substitution of a small pad of Pyrex glass wool for the frit will
         prevent cross contamination of sample extracts).

5.2   Kuderna-Danish (K-D) Apparatus

5.2.1    Concentrator Tubes - 15 mL and 10 mL graduated (Kontes K-570050-1025
         and 570040-1025 or equivalent).  Calibrations must be checked at the
         volumes employed in the test.   Ground-glass stoppers are used to
         prevent evaporation of extracts.

5.2.2    Evaporative Flasks - 500 mL (Kontes K-570001-0500 or equivalent).
         Attach to concentrator tube with springs.

5.2.3    Snyder Column - Two-ball Micro (Kontes K-569001-0219 or equivalent).

5.3   Spatula - stainless steel or Teflon
                                   D-72/SVOA                            OLM03.0

-------
                                          Exhibit  D Semivolatiles  —  Appendix A
                                     Screening of  Semivolatile  Organic  Extracts
 5.4    Balances  -  Analytical,  capable of accurately weighing ± 0.0001  g and one
       capable of  weighing  100 g  to  ± 0.01  g.   The balances  must  be calibrated
       with  class  S  weights or known reference weights  once  per each 12-hour
       work  shift.   The  balances  must be calibrated with class  S  weights at a
       minimum of  once per  month.  The balances must also be annually  checked
       by  a  certified technician.

 5.5    Ultrasonic  Cell Disrupters -  Heat Systems,  Ultrasonic Inc.,  Model W-385
       Sonicator (475 watt  with pulsing capability,  No.  200,  1/2  inch  tapped
       disrupter horn, No.  419 1/8 inch standard tapered Microtip probe,  and
       No. 305,  3/4  inch tapered  high gain,  1Q'  disrupter horn, or  No.  208  3/4
       inch  standard solid  disruptor horn),  or equivalent devices with a
       minimum of  375 watt  output capability.

       NOTE:  In order to ensure  that sufficient energy  is transferred to the
       sample during extraction,  the Microtip  probe  or horn  must  be replaced if
       the tip begins to erode.   Erosion of  the tip  is evidenced  by a  rough
       surface.

 5.6    Sonabox Acoustic  Enclosure  -  recommended with above disrupters  for
       decreasing  cavitation sound.

 5.7    Vacuum Filtration Apparatus

 5.7.1    Buchner  Funnel

 5.7.2    Filter paper -  Whatmann  No.  41  or  equivalent

 5.8    Pyrex Glass Wool  - rinsed with methylene  chloride.

 5.9    Silicon Carbide Boiling Chips  -  approximately 10/40 mesh.  Heat  to 400
       °C for 30 minutes  or Soxhlet  extract  with methylene chloride.

 5.10   Water Bath  -  heated, with  concentric  ring cover,  capable of  temperature
       control (±  2  °C).   The  bath should be used  in a hood.

 5.11   Nitrogen Evaporation Device -  equipped  with a water bath that can  be
       maintained  at 35  - 40 °C.   (N-Evap by Organomation  Associates,  Inc.,
       South Berlin, MA,   or eg_uivalent) .  To prevent the  release  of solvent
       fumes into the laboratory, the nitrogen evaporation device must  be used
       in a hood.

 5.12   Gas Chromatograph/Flame  lonization Detector  (GC/FID System)

 5.12.1   Gas Chromatograph -  an analytical  system complete with  a  temperature
         programmable gas Chromatograph and all required accessories  including
         syringes,  analytical columns, and  gases.   The  injection port  must  be
         designed for on-column injection when  using packed columns and  for
         splitless  injection when using capillary columns.

5.12.2   Gas Chromatograph Column -  30 m (or  longer) X 0.32 mm,  1  micron film
         thickness, silicone  coated, fused  silica capillary column  (J  &  W
         scientific DB-5 or equivalent).  Note  that this is minimum
         requirement for column length.  Longer columns may be used.

                                  D-73/SVOA                           OLM03.0

-------
 Exhibit D  Semivolatiles — Appendix A
 Screening  of  Semivolatile Organic Extracts
 5.12.3   Flame  lonization Detector

 6.0   REAGENTS  AND STANDARDS

 6.1   Reagents

 6.1.1    Reagent Water - defined as water in which an interferant is not
         observed at or above the CRQL for each analyte of interest.

 6.1.2    Sodium Thiosulfate - (ACS) granular.

 6.1.3    Sulfuric Acid Solution (1+1) - slowly add 50 mL of concentrated H2S04
         (sp.gr. 1.84) to 50 mL of reagent water.

 6.1.4    Acetone, methanol, methylene chloride - pesticide residue analysis
         grade  or equivalent.

 6.1.5    Sodium Sulfate - anhydrous powdered reagent grade.  Purify by heating
         at 400 °C for four hours in a shallow tray.  Cool in a desiccator and
         store  in a glass bottle (Baker anhydrous powder, catalog #73898 or
         equivalent).

 6.2   Standards

 6.2.1    Introduction

         The contractor must provide all standard solutions to be used with
         this contract.  These standards may be used only after they have been
         certified according to the procedure in Exhibit E.  The contractor
         must be able to verify that the standards are certified.
         Manufacturer's certificates of analysis must be retained by the
         contractor and presented upon request.

6.2.2    Stock Standard Solution

6.2.2.1     Prepare or purchase a stock standard solution containing phenol,
            phenanthrene,  and di-n-octyl phthalate at concentrations of 1
            ng/l±L.  Prepare stock standard solutions by accurately weighing
            about 0.01 g of pure material.   Dissolve the material in pesticide
            quality methylene chloride and dilute to volume in a 10 mL
            volumetric flask.  Larger volumes may be used at the convenience
            of the analyst.   If compound purity is assayed at 97.0 percent or
            greater,  the weight may be used without correction to calculate
            the concentration of the stock standard.  If the compound purity
            is assayed to be less than 97.0 percent, the weight must be
            corrected when calculating the concentration (see Exhibit E,
            Analytical Standards Requirements).  Commercially prepared stock
            standards may be used at any concentration if they are certified
            by the manufacturer or by an independent source (see Exhibit E).

6.2.2.2     Transfer  the stock standard solutions into Teflon-sealed screw-cap
            bottles.   Stock standard solutions should be checked frequently
            for signs of degradation or evaporation, especially just prior to
            preparing calibration standards from them.  Stock standard

                                   D-74/SVOA                            OLM03.1

-------
                                         Exhibit D  Semivolatiles  — Appendix A
                                     Screening  of Semivolatile Organic Extracts


             solutions must be  replaced  after six months,  or  sooner  if
             comparison with  quality  control check samples indicates a problem.

6.2.3    Working Standard Solutions

6.2.3.1      Surrogate Standard Spiking  Solution

             Prepare  a surrogate  standard spiking solution that  contains
             nitrobenzene-ds,  terphenyl-d^,  2-fluorobiphenyl,  and  1,2-
             dichlorobenzene-d^, at a concentration of 100 /xg/mL;  phenol-d5,
             2,4,6-tribromophenol, 2-fluorophenol, and  2-chlorophenol-d4 at a
             concentration of 150 /tg/mL.  Surrogate  standards are  added to all
             samples  and calibration  solutions.  Additional surrogates may be
             added at the laboratory's discretion.

6.2.3.2      GC Calibration Standard

             Prepare  a working  standard mixture of phenol, phenanthrene and di-
             n-octylphthalate.  The concentration must  be  such that the volume
             injected equals  50 ng of each compound.

6.2.4    Storage of  Standards

6.2.4.1      Store the stock  standard solutions at less than 4 °C  but not
             greater  than 6 °C  in Teflon-lined  screw-cap amber bottles and
             protect  from light.

6.2.4.2      Store the working  standards at less than 4 °C but not greater than
             6 °C in  Teflon-sealed containers.  The  solutions should be checked
             frequently for stability.  These solutions must be  replaced after
             six months or sooner if comparison with quality control check
             samples  indicates  a problem.

6.2.4.3      Samples, sample extracts and standards must be stored separately.

7.0   QUALITY CONTROL

7.1   Method Blank

7.1.1    Summary

         A method blank is a volume of a clean reference matrix ( reagent
         water for water samples, or purified sodium sulfate for  soil/sediment
         samples)  that is carried through the entire analytical procedure.
         The volume  or weight of the reference matrix must be approximately
         equal to the volume or weight of samples associated with the blank.
         The purpose of a method blank is to determine the levels of
         contamination associated with the processing and analysis of samples.

7.1.2    Frequency

         One method  blank must be extracted and analyzed on each GC/FID system
         used to screen samples for the following,  whichever is most frequent.
                                   D-75/SVOA                           OLM03.0

-------
 Exhibic D Semivolatiles — Appendix A
 Screening of Semivolatile Organic Extracts
          •   Each  SDG,  or

          •   Each 20  samples  in a   SDG,  including matrix  spike  and re-
              analyses, that are of  similar matrix  (water,  soil,  or sediment)
              or  similar  concentration  (soil only), or

          •   Whenever samples are extracted by the same procedure  (continuous
              liquid-liquid extraction  or  sonication) .

 7.1.3     Procedure

          For  semivolatile  analyses, a  method blank for water samples  consists
          of a 1  L volume of reagent water spiked with 0.5  mL of  the surrogate
          spiking solution.  For medium and low level soil/sediment samples,
          method  blanks consist  of 1 g  and 30 g of sodium  sulfate spiked with
          0.5  mL  of surrogate  spiking solution, respectively.  Extract and
          concentrate  method blanks at  the same time as the samples associated
          with the blanks according to  Section 12.1. Analyze the  method blank
          according to Section  12.1.

 8.0   CALIBRATION AND STANDARDIZATION

 8.1   GC/FID  Operating Conditions

      Suggested GC  operating conditions are as follows:

      •   Initial  column temperature hold - 20 °C for 4 minutes

      •   Column  temperature program - 20-280 °C at 8 °C /minute

      •   Final column temperature hold -280 °C for 8 minutes

      •   Injector  - Grob-type,  splitless

      •   Sample volume  - 1-2 /tL

      •   Carrier gas - Helium at. 30 cm/sec

8.2   GC Calibration

8.2.1    Summary

         Prior to sample analysis,  each GC/FID system must be initially
         calibrated at one concentration level to determine instrument
         sensitivity.

8.2.2    Frequency

         Each GC/FID system must be calibrated at the beginning of each 12-
         hour shift.
                                  D-76/SVOA                            OLM03.0

-------
                                         Exhibit D Semivolatiles — Appendix A
                                    Screening of Semivolatile Organic Extracts
8.2.3    Procedure
         Inject 1-2 juL of the GC calibration standard prepared in Section
         6.2.3.2.  The volume injected must equal 50 ng of each of the
         calibration compounds.

8.2.4    Technical Acceptance Criteria

8.2.4.1     The GC must be standardized for half-scale response from 50 ng of
            phenanthrene.

8.2.4.2     The GC must adequately separate phenol from the solvent front.

8.2.4.3     A minimum of quarter-scale response for 50 ng of di-n-
            octylphthalate must be exhibited.

8.2.5    Corrective Action

8.2.5.1     If the technical acceptance criteria are not met, recalibrate the
            GC instrument.  It may be necessary to change the column or take
            other corrective actions to achieve the acceptance criteria.

8.2.5.2     GC calibration technical acceptance criteria must be met before
            any samples are injected.

9.0   PROCEDURE

9.1   Sample Preparation

9.1.1    Low Level Soil/Sediment

9.1.1.1     Decant and discard any water layer on a sediment sample.  Mix
            samples thoroughly, especially composited samples. Discard any
            foreign objects such as sticks leaves and rocks.

9.1.1.2     Weigh approximately 30 g of sample to the nearest 0.1 g into a 400
            mL beaker and proceed with low level soil/sediment sample
            preparation as described in Section 10.1.4.4 of the Semivolatile
            Analytical Method.

9.1.1.3     Take 5 mL from the 300 mL (approximate) total extract and
            concentrate to 1 mL following Section 10.2.2.1 or 10.2.2.2 of
            Exhibit D (Semivolatiles), but note that the final volume for
            screening is 1 mL, not 0.5 mL.

9.1.2    Medium Level Soil/Sediment

9.1.2.1     Decant and discard any water layer on a sediment sample.  Mix
            samples thoroughly, especially composited samples.  Discard any
            foreign objects such as sticks, leaves and rocks.

9.1.2.2     Transfer approximately 1 g (record weight to the nearest 0.1 g) of
            sample to a 20 mL vial.  Wipe the mouth of the vial with a tissue
            to remove any sample material and proceed with the medium level

                                   D-77/SVOA                            OLM03.0

-------
Exhibit D Semivolatiles — Appendix A
Screening of Semivolatile Organic Extracts


            sample preparation procedure described in Section 10.1.4.5 of
            Exhibit D (Semivolatiles).

9.1.2.3     Take 5 mL from the 10 mL total extract and concentrate to 1 mL
            following Section 10.2.2.1 or 10.2.2.2 of Exhibit D
            (Semivolatiles).

9.2   GC/FID Analysis

      Inject 1-2 fjiL of extract.

9.3   Interpretation of Chromatograms

9.3.1    Soil/Sediment

9.3.1.1     If no sample peaks from the extract (from low or medium level
            preparation) are detected, or all are less than 10.0 percent full
            scale deflection, the sample must be prepared by the low level
            protocol.

9.3.1.2     Peaks are detected at greater than 10.0 percent full scale
            deflection and less than or equal to full scale deflection.

            •   If the screen is from the medium level extract,  proceed with
                GC/MS analysis of this extract with appropriate dilution if
                necessary.

            •   If the screen is from the low level extract, discard the
                extract and prepare the samples by medium level method for
                GC/MS analysis.

9.3.1.3     Peaks are detected at greater than full scale deflection.

            •   If the screen is from the medium level preparation, calculate
                the dilution necessary to reduce the major peaks to between
                half and full scale deflection.  Use this dilution factor to
                dilute the extract.  This dilution is analyzed by GC/MS for
                extractable organics.

            •   If the screen is from the low level preparation, discard the
                extract and prepare the samples by the medium level method for
                GC/MS analysis.
                                   D-78/SVOA                            OLM03.0

-------
       EXHIBIT D
  ANALYTICAL METHODS
FOR PESTICIDES/AROCLORS
       D-l/PEST                            OLM03.0

-------
            Exhibit D - Analytical Methods for Pesticides/Aroclors

                               Table of Contents


Section                                                                   Page

1.0   SCOPE AND APPLICATION  	    4

2.0   SUMMARY OF METHOD	    5
      2.1   Water	    5
      2.2   Soil/Sediment  	    5

3.0   DEFINITIONS	    5

4.0   INTERFERENCES 	    6

5.0   SAFETY	    6

6.0   EQUIPMENT AND SUPPLIES  	    7

7.0   REAGENTS AND STANDARDS	   14
      7.1   Reagents	   14
      7.2   Standards	   14
      7.3   Storage of Standard Solutions .	   18

8.0   SAMPLE COLLECTION, PRESERVATION AND STORAGE 	   20
      3.1   Sample Collection and Preservation  	   20
      8.2   Procedure for Sample Storage  	   20
      8.3   Procedure for Sample Extract Storage  	   20
      8.4   Contract Required Holding Times  	   20

9.0   CALIBRATION AND STANDARDIZATION 	   22
      9.1   Gas Chromatograph Operating Conditions  	   22
      9.2   Initial Calibration 	   22
      9.3   Calibration Verification  	   30

10.0  PROCEDURE	   35
      10.1  Sample Preparation  	   35
      10.2  GC/EC Analysis	   57

11.0  DATA ANALYSIS AND CALCULATIONS	   62
      11.1  Qualitative Identification  	   62
      11.2  Calculations	   65
      11.3  Technical Acceptance Criteria for Sample Analysis 	   69
      11.4  Corrective Action for Sample Analysis 	   71

12.0  QUALITY CONTROL	   72
      12.1  Blank Analyses	   72
      12.2  Matrix Spike/Matrix Spike Duplicate (MS/MSD)  	   78

13.0  METHOD PERFORMANCE  	   82

14.0  POLLUTION PREVENTION   	   82


                                   D-2/PEST                             OLM03.0

-------
15.0  WASTE MANAGEMENT   	    82





16.0  REFERENCES	    82





17.0  TABLES/DIAGRAMS/FLOWCHARTS   	    83
                                   D-3/PEST                             OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 1
Scope and Application
1.0   SCOPE AND APPLICATION

1.1   The analytical method that follows is designed to analyze water,
      sediment and soil from hazardous waste sites to determine the presence
      and concentration of the chlorinated pesticides and Aroclors found in
      the Target Compound List (TCL)  in Exhibit C.  The method can be used for
      determining analyte concentrations in the range from the contract
      required quantitation limits (CRQL)  to one million times the CRQL in
      these matrices when appropriate dilutions are made.  The method is based
      on EPA Method 608, and it covers sample extraction, extract cleanup
      techniques and GC/EC analytical methods for pesticides and Aroclors.

1.2   Resolution difficulties have been associated with the following pairs of
      compounds using this method:

      •   On a DB-608 or equivalent column, DDE and dieldrin; methoxychlor and
          endrin ketone; and endosulfan I  and gamma-Chlordane.

      •   On a DB-1701 or equivalent  column, endosulfan I and gamma-Chlordane,
          and methoxychlor and endosulfan  sulfate.

1.3   There are two isomers of heptachlor  epoxide, the endo isomer (isomer A)
      and the exo isomer (isomer B).   The  two isomers are separable using
      current GC capillary columns.  Only  the exo isomer (isomer B) is of
      environmental significance.  This is the isomer that must be used as an
      analytical standard,  identified and  quantitated in sample analysis, and
      reported on appropriate forms as heptachlor epoxide.
                                   D-4/PEST                            OLM03.0

-------
                               Exhibit D Pesticides/Aroclors — Sections 2 & 3
                                                 Summary of Method/Definitions
2.0   SUMMARY OF METHOD

2.1   Water

      Continuous liquid-liquid or separatory funnel extraction procedures are
      employed for aqueous samples.  AIL volume of sample is spiked with the
      surrogate solution and extracted with methylene chloride using a
      separatory funnel or a continuous extractor.  The methylene chloride
      extract is dried with anhydrous sodium sulfate, concentrated and cleaned
      up by GPC (GPC is required when higher molecular weight compounds are
      present that interfere with the analyses of target compounds; GPC is
      optional for all other circumstances).  The extract is then solvent-
      exchanged into hexane, cleaned up by Florisil cartridges, and the final
      volume adjusted to 1 mL or 2 mL.  The extract is analyzed using a dual
      column wide-bore capillary Gas Chromatography/Electron Capture (GC/EC)
      technique.

2.2   Soil/Sediment

      A 30 g aliquot of sample is spiked with the surrogate and then mixed
      with anhydrous sodium sulfate and extracted with a 1:1 acetone/methylene
      chloride solvent mixture by sonication.  The extract is filtered,
      concentrated and solvent-exchanged into methylene chloride.  The
      methylene chloride extract is then cleaned up by GPC (mandatory),
      solvent-exchanged into hexane, cleaned up by Florisil cartridge,  and
      adjusted to a final volume of 1 mL or 2 mL.  The extract is analyzed
      using a dual column wide-bore capillary Gas Chromatography/Electron
      Capture (GC/EC)  technique.

3.0   DEFINITIONS

      See Exhibit G for a complete list of definitions.
                                   D-5/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Sections 4 & 5
Interferences/Safety
4.0   INTERFERENCES

4.1   Method interferences may be caused by contaminants in solvents,
      reagents, glassware, and other sample processing hardware.  These
      contaminants lead to discrete artifacts or to elevated baselines in gas
      chromatograms.  Routinely, all of these materials must be demonstrated
      to be free from interferences under the conditions of the analysis by
      running reagent blanks.  Interferences caused by phthalate esters can
      pose a major problem in pesticide analysis.  Common flexible plastics
      contain varying amounts of phthalates which are easily extracted during
      laboratory operations, so cross-contamination of glassware frequently
      occurs when plastics are handled.  Interferences from phthalates can
      best be minimized by avoiding the use of such plastics in the
      laboratory.

4.2   Matrix interferences may be caused by contaminants that are co-extracted
      from the sample.  The extent of matrix interferences will vary
      considerably from source to source, depending upon the nature of the
      site being sampled.   The cleanup procedures must be used to remove such
      interferences in order to achieve the contract required quantitation
      limits.

5.0   SAFETY

5.1   The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely determined; however,  each chemical should be treated
      as a potential health hazard.   Exposure to these rtagents should be
      reduced to the lowest possible level.  The laboratory is responsible for
      maintaining a current awareness file of OSHA regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
      file of data handling sheets should be made available to all personnel
      involved in these analyses.   Specifically, concentrated sulfuric acid
      and the 10 N sodium hydroxide solution are moderately toxic and
      extremely irritating to skin and mucous membranes.   Use these reagents
      in a fume hood whenever possible and if eye or skin contact occurs,
      flush with large volumes of  water.  Always wear safety glasses or a
      shield for eye protection,  protective clothing, and observe proper
      mixing when working with these reagents.

5.2   The following analytes covered by this method have been tentatively
      classified as known or suspected human or mammalian carcinogens: 4,4'-
      DDT,  4,4'-DDD,  the BHCs, and the Aroclors.  Primary standards of these
      toxic compounds should be prepared in a hood.  A NIOSH/MESA approved
      toxic gas respirator should  be worn when the analyst handles high
      concentrations of these toxic compounds.
                                   D-6/PEST                            OLM03.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 6
                                                        Equipment and Supplies
6.0   EQUIPMENT AND SUPPLIES

      Brand names, suppliers, and part numbers are for illustrative purposes
      only.  No endorsement is implied.  Equivalent performance may be
      achieved using equipment and supplies other than those specified here,
      but demonstration of equivalent performance meeting the requirements of
      this Statement of Work is the responsibility of the Contractor.  The
      Contractor must document in its Narrative when it uses equipment and
      supplies other than those specified here.

6.1   Glassware

6.1.1    Continuous Liquid-Liquid Extractors - equipped with Teflon or glass
         connecting joints and stopcocks requiring no lubrication (Hershberg-
         Wolf extractor. Ace Glass Company, Vineland, NJ P/N 6841-10 or
         equivalent) or Hydrophobic Membrane-based Extractor (Accelerated One
         Step™ Extractor, Corning series 3195 or equivalent).

6.1.2    Separatory Funnels - 2 L with Teflon stopcock.

6.1.3    Beakers - 400 mL.

6.1.4    Erlenmeyer Flasks - 250 mL.

6.1.5    Syringes - 10 mL with Luerlok fitting, 1 mL or 2 mL.

6.1.6    Vials and Caps - 20 mL and 10 mL (optional) with screw cap and Teflon
         or aluminum foil liner, 2 mL capacity for GC auto sampler.

6.1.7    Pipets - glass volumetric 1 mL or 2 mL.

6.1.8    Centrifuge Tube - 12 to 15 mL with 19 mm ground glass joint
         (optional).

6.1.9    Graduated Cylinder - 1 L capacity.

6.1.10   Drying Column - chromatogiaphic column approximately 400 mm long x 19
         mm ID, with coarse frit.   (Substitution of a small pad of disposable
         Pyrex glass wool for the frit will help prevent cross-contamination
         of sample extracts).

6.1.11   Volumetric Flasks - 10 mL and 1 or 2 mL.

6.1.12   Bottle or Test Tube - 20 mL with Teflon-lined screw cap for sulfur
         removal and a glass bottle - 1 L volume, for use in preparation of
         Bio Beads for packing into column.

6.1.13   Powder Funnels - 10 cm diameter, for filtration/drying.

6.1.14   Buchner Funnels - 9 cm diameter, for filtration.
                                   D-7/PEST                            OLM03.0

-------
Exhibit D  Pesticides/Aroclors — Section 6
Equipment  and Supplies
 6.2   Kuderna-Danish  (K-D) Apparatus.

 6.2.1    Concentrator Tubes - 10 mL, graduated  (Kontes K-570040-1029, or
         equivalent).

 6.2.2    Evaporative Flasks - 500 mL (Kontes K-470001-0500, or equivalent).

 6.2.3    Snyder Column - three-ball macro  (Kontes K-503000-0121, or
         equivalent).

 6.2.4    Snyder Column - two-ball micro  (Kontes K-569001 -0219, or
         equivalent).

 6.3   Vacuum System for Eluting Multiple Cleanup Cartridges.

 6.3.1    Vac Elute Manifold - Analytichem  International, J.T. Baker, or
         Supelco (or equivalent).  The manifold design must ensure that there
         is no contact between plastics containing phthalates and sample
         extracts.

 6.3.2    Vacuum Trap - made from a 500 mL  sidearm flask fitted with a one-hole
         stopper and glass tubing.

 6.3.3.   Vacuum Pressure Gauge.

 6.3.4    Rack for holding 10 mL volumetric flasks in the  manifold.

      NOTE:  Ot >r types of equivalent systems,  .such as an automated system
      using syringe pressure,  are considered to be acceptable for elution of
      florisil cartridges,  as long as all QC and sample technical acceptance
      criteria are met.

 6.4   pH Paper - wide range (Hydrion Papers, Micro-essential Laboratory,
      Brooklyn, NY, or equivalent).

 6.5   Spatula - stainless steel or Teflon.

 6.6   Centrifuge -  table top (optional).

6.7   Balances - top loading,  capable of weighing accurately to + 0.01 g,
      analytical,  capable of weighing accurately to ± 0.0001 g.  The balances
      must be calibrated with class  S weights or known reference weights once
      per each 12-hour work shift.  The balances must be calibrated with class
      S  weights at  a minimum of once per month.   The balances must also be
      annually checked by a certified technician.

6.8   Ultrasonic Cell  Disrupter - Heat Systems,  Ultrasonics, Inc., Model W-385
      (475 watt with pulsing capability,  No. 207 3/4-inch tapered disruptor
      horn) or equivalent device with a minimum 375 watt output capability.
      NOTE:  In order  to ensure that sufficient energy is transferred to the
      sample during extraction,  the  horn must be replaced if the tip begins to
      erode.  Erosion  of the tip is  evidenced by a rough surface.
                                   D-8/PEST                            OLM03.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 6
                                                        Equipment and Supplies
6.9   Sonabox Acoustic Enclosure (or equivalent)  - for use with disrupter to
      decrease noise level.

6.10  Filter Paper - No. 41 Whatmann (or equivalent),  9 cm circles (optional).

6.11  Pyrex Glass Wool - rinsed with methylene chloride and dried before use.

6.12  Boiling chips.

6.12.1   Silicon carbide boiling chips - approximately 10 to 40 mesh.   Heat
         the chips to 400 °C for 30 minutes or solvent rinse before use.

6.12.2   Teflon boiling chips (optional) -  solvent rinse the chips before
         use.

6.13  Water Bath - heated, with concentric ring cover, capable of temperature
      control.  NOTE:  The water bath should be used in a hood.

6.14  GPC Cleanup System

6.14.1   Gel Permeation Chromatography System - GPC Autoprep Model 1002 A or
         B, Analytical Biochemical Laboratories,  Inc., or equivalent.   Systems
         that perform satisfactorily have been assembled from the following
         components: an HPLC pump, an auto sampler or a valving system with
         sample loops, and a fraction collector.   All systems, whether
         automated or manual, must meet the calibration requirements of
         Section 10.1.8.1.  NOTE: GPC cleanup is required for extracts for all
         soils/sediments and for water extracts containing higher molecular
         weight contaminants that interfere with the analyses of the target
         compounds.

6.14.1.1    Chromatographic column - 700 mm x 25 mm ID glass column.  Flow is
            upward.  To simplify switching from the UV detector during
            calibration to the GPC collection device during extract cleanup,
            an optional double 3-way valve (Rheodyne Type 50 Teflon Rotary
            Valve #10-262 or equivalent) may be attached so that the column
            exit flow can be shunted either to the UV flow-through cell or to
            the GPC collection device.

6.14.1.2    Guard column  (optional) - 5 cm, with appropriate fittings to
            connect to the inlet side of the analytical column (Supelco 5-8319
            or equivalent).

6.14.1.3    Bio Beads (S-X3) - 200 to 400 mesh, 70 g  (Bio-Rad Laboratories,
            Richmond, CA, Catalog 152-2750 or equivalent).  An additional 5 g
            of Bio Beads is  required if the optional guard column is
            employed.  The quality of Bio Beads may vary from lot to lot
            because of excessive fines in some lots.  In addition to fines
            having a detrimental effect on chromatography, they can also pass
            through the column screens and damage the valve.

6.14.1.4    Ultraviolet detector - fixed wavelength (254 nm) with a semi-prep
            flow-through cell.
                                   D-9/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 6
Equipment and Supplies
6.14.1.5    Strip chart recorder - recording integrator or laboratory data
            system.

6.14.1.6    Syringe filter assembly, disposable - Bio-Rad "Prep Disc" sample
            filter assembly #343-0005, 25 mm, and 5 micron filter discs or
            equivalent.  Note:  Some instrument manufacturer's recommend a
            smaller micron filter disc.  Consult your instrument operation
            manual to determine the proper size filter disc to use in your
            system.  Check each batch for contaminants.  Rinse each filter
            assembly (prior to use) with methylene chloride if necessary.

6.15  Florisil - 500 mg or 1 g cartridges with stainless steel or Teflon
      frits, (Catalog No. 694-313,  Analytichem, 24201 Frampton Ave., Harbor
      City, CA, or equivalent).

6.16  Nitrogen Evaporation Device - equipped with a heated bath that can be
      maintained at 35 to 40 °C (N-Evap by Organomation Associates, Inc.,
      South Berlin, MA, or equivalent).

6.17  Oven - drying.

6.18  Desiccator.

6.19        Crucibles - porcelain crucibles or aluminum weighing pans.

6.20        pH Meter - with a combination glass electrode.  Calibrate
            according to manufacturer's instructions.  pH meter must be
            calibrated prior to each use.

6.21        Magnetic Stirrer Motor  - Model PC 353,  Corning Co., Corning, NY,
            or equivalent.

6.22        Magnetic Stirrer Bar -  Teflon coated,  at least 4 cm long.

6.23        Gas Chromatograph/Electron Capture Detector (GC/EC) System.
6.23.1   Gas Chromatograph - must adequately regulate temperature in order to
         give a reproducible temperature program and have a flow controller
         that maintains a constant  column flow rate throughout temperature
         program operations.   The system must be suitable for splitless
         injection and have all required accessories including syringes,
         analytical columns and gases.

6.23.2   Gas chromatographs that are available from some manufacturers may
         have difficulty in meeting certain method QC requirements because of
         endrin and DDT breakdown in the injector.   This problem can be
         minimized by operating the injector at 200 - 205 °C,   using a Pvrex
         (not quartz)  methyl silicone deactivated injector liner, and
         deactivating the metal parts in the injector with dichlorodimethyl
         silane.  In some cases, using a 0.25 inch packed column injector
         converted for use with 0.53 mm capillary columns works better than a
         Grob-type injector.   If a  Grob-type injector is used, a 4 mm liner
         may be required to meet breakdown criteria.
                                  D-10/PEST                            OLM03.0

-------
                                     Exhibit  D  Pesticides/Aroclors  —  Section  6
                                                         Equipment  and Supplies
 6.23.3    Gas  Chromatograph  Columns  -  two  wide-bore  (0.53  mm ID)  fused silica
          GC columns  are  required.   A  separate  detector  is required  for each
          column.   The  specified  analytical  columns  are  a  30 m x
          0.53 mm  ID, 1.0 /xm film thickness  DB-1701  (J&W Scientific);  SPB  1701
          (Supelco);  AT 1701 (Alltech); RTX-1701  (Restek); CP-Sil  19CB
          (Chrompack);  007-1701  (Quadrex); BP-10  (SGE);  or equivalent,  and a  30
          m x  0.53  mm ID, 0.5 to  1.0 /un film thickness DB-608 (J&W Scientific);
          HP-608  (Hewlett Packard);  SPB-608  (Supelco); 007-608 (Quadrex);  BP-
          608  (SGE);  CP-Sil  8CB  (Chrompack); or equivalent.   NOTE:   The column
          length stated above is  the minimum requirement.   Longer  columns  that
          meet resolution and calibration  requirements may be used.  A
          description of  the GC columns used for  analysis  shall be provided in
          the  SDG narrative.

 6.23.3.1    PACKED COLUMNS  CANNOT BE  USED.

 6.23.3.2    A capillary  column is considered equivalent if:

            •   The  column  does  not introduce  contaminants  which  interfere
                with identification and quantitation of the compounds  listed
                in Exhibit  C (Pesticides).

            •   The  analytical results generated using the  column meet  the
                initial  calibration and calibration verification  technical
                acceptance  criteria listed  in  the SOW and the CRQLs listed in
                Exhibit  C (Pesticides).

            •   The  column  can accept at  least 16 times the low point  standard
                for  individual standard mixtures A and B  for  each compound
                listed in Exhibit C (Pesticides) without  becoming overloaded.

            •   The column  pair  chosen must have dissimilar phases/chemical
                properties  in order to separate the compounds of  interest  in
                different RT order.

6.23.3.3    Although the instructions included in the SOW are for wide  bore
            capillary columns,  narrower bore capillary columns may be
            evaluated for use.

6.23.3.4    As applicable,  follow the manufacturer's instructions for use  of
            its product.

6.23.3.5    The Contractor must maintain documentation that the column  met the
            criteria in 6.23.3.2.   The minimum documentation  is as follows:

6.23.3.5.1     Manufacturer provided information concerning the performance
               characteristics of the column;

6.23.3.5.2     Chromatograms and data system reports generated on the GC/ECD
               and used for CLP analyses:
                                   D-11/PEST                            OLM03.1

-------
Exhibit D Pesticides/Aroclors — Section 6
Equipment and Supplies
               •   From instrument blanks which demonstrate that there are no
                   contaminants which interfere with the pesticide analysis
                   when using the alternate column;

               •   For initial calibration standards analyzed using the
                   column;

               •   For calibration verification standards analyzed using the
                   alternate column.

6.23.3.5.3     Based on the Contractor generated data described in 6.23.3.5.2,
               the Contractor must complete a written review, signed by the
               Laboratory Manager certifying that:

               •   The column performance is comparable to the required column
                   performance in its ability to produce initial calibration
                   and calibration verifications which meet the technical
                   acceptance criteria in 9.2.5 and 9.3.5.

               •   The low point initial calibration standard analyses have
                   adequate sensitivity to meet the pesticide CRQLs.

               •   The high point initial calibration standard analyses were
                   not overloaded.

               •   The column does not introduce contaminants which interfere
                   with identification and quantitation of compounds listed in
                   Exhibit C (Pesticides).

6.23.3.5.4     The documentation must be made available to the Agency during
               on-site laboratory evaluations or sent to the Agency upon
               request of the Technical Project Officer or the Administrative
               Project Officer.

6.23.3.6    Columns are mounted in a 0.25-inch injector ports by using glass
            adapters available from a variety of commercial sources (J&W
            Scientific, Supelco,  Inc., Hewlett-Packard, Varian, Inc.,  Perkin
            Elmer, or equivalent).   The two columns may be mounted into a
            single injection port with a tee adapter (Supelco, Inc.,
            Bellefonte, PA, Catalog No. 2-3660, or equivalent).  Use of this
            adapter allows simultaneous injection onto both columns.  The
            laboratory should follow manufacturer's recommendations for
            mounting 0.53 mm capillary columns in injector ports.

6.23.3.7    The carrier gas for routine applications is helium.  Laboratories
            may choose to use hydrogen as a carrier gas, but they must clearly
            identify its use in the SDG Narrative and on all divider pages
            preceding raw chromatographic data in submissions to the Agency.
            Laboratories that choose to use hydrogen are advised to exercise
            caution in its use.   Use of a hydrogen leak detector is highly
            recommended when hydrogen is used as the carrier gas.  All GC
            carrier gas lines must be constructed from stainless steel or
            copper tubing.  Non-polytetrafluoroethylene (PTFE) thread sealants
            or flow controllers with rubber components are not to be used.

                                  D-12/PEST                            OLM03.1

-------
                                    Exhibit D Pesticides/Aroclors — Section 6
                                                        Equipment and Supplies
6.23.4   Electron Capture Detector (ECD) - the linearity of the response of
         the ECD may be greatly dependent on the flow rate of the make-up gas.
         The make-up gas must be P-5, P-10 (argon/methane) or nitrogen
         according to the instrument specification.   Care must be taken to
         maintain stable and appropriate flow of make-up gas to the detector.
         The GC/EC system must be in a room in which the atmosphere has been
         demonstrated to be free of all contaminants which may interfere with
         the analysis.  The instrument must be vented to outside the facility
         or to a trapping system which prevents the  release of contaminants
         into the instrument room.

6.23.5   Data System - a data system must be interfaced to the GC/EC.   The
         data system must allow the continuous acquisition of data throughout
         the duration of the chromatographic program and must permit,  at the
         minimum, the output of time vs. intensity (peak height or peak area)
         data.  Also, the data system must be able to rescale chromatographic
         data in order to report chromatograms meeting the requirements listed
         within this method.
                                   D-13/PEST                             OLM3.0

-------
 Exhibit  D  Pesticides/Aroclors — Section 7
 Reagents and  Standards
 7.0   REAGENTS AND STANDARDS

 7.1   Reagents

 7.1.1    Reagent water - defined as water in which no interferent is observed
         at one-half the CRQL of any pesticide/Aroclor when one liter of the
         reagent water is extracted and prepared by using the same workup
         procedure as for a water sample.

 7.1.2    Sodium sulfate - granular-anhydrous reagent grade, heated at 400  °C
         for 4 hours, or at 120 °C for 16 hours, cooled in a desiccator, and
         stored in a glass bottle.  Each lot must be extracted with hexane and
         analyzed by GC/EC to demonstrate that it is free of interference
         before use.  Baker anhydrous granular, Catalog No. 3375, or
         equivalent.  CAUTION:  An open container of sodium sulfate may become
         contaminated during storage in the laboratory.

 7.1.3    Concentrated sulfuric acid (H2SC>4)-  18  N.

 7.1.4    Sodium hydroxide solution (NaOH) (10 N) -  carefully dissolve 40  g of
         NaOH in reagent water and dilute the solution to 100 mL.

 7.1.5    10 percent acetone in hexane (v/v)  -  prepare by adding 10 mL of
         acetone to 90 mL of hexane.   NOTE:   Prepare this mixture accurately
         or the results from the Florisil cartridge cleanup will be adversely
         affected.   Water in the acetone also will adversely affect Florisil
         performance.

 7.1.6    Methylene chloride, hexane,  acetone, toluene,  iso-octane, and
         methanol (optional) - pesticide quality or equivalent.   It is
         recommended that each lot of solvent used be analyzed to demonstrate
         that it is free of interference before use.   Methylene chloride must
         be certified as acid free or must be tested to demonstrate that it is
         free of hydrochloric acid.   Acidic methylene chloride must be passed
         through basic alumina and then demonstrated to be free of
         hydrochloric acid.

7.1.7    Mercury -  triple distilled,  for sulfur cleanup.

7.1.8    Copper powder (optional)  -  fine, granular (Mallinckrodt ^649 or
         equivalent).  Copper may  be  used instead of mercury for sulfur
         cleanup.   Remove oxides by  treating with dilute nitric acid, rinse
         with distilled water to remove all traces of acid, rinse with
         acetone,  and dry under a  stream of nitrogen.

7.2   Standards

7.2.1    Introduction

         The Contractor must provide  all standards to be used with this
         contract.   These standards  may be used only after they have been
         certified  according to the  procedure in Exhibit E.  The Contractor
         must be able to verify that  the standards are certified.
                                  D-14/PEST                             OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 7
                                                        Reagents and Standards
         Manufacturer's certificates of analysis must be retained by the
         Contractor and presented upon request.

7.2.2    Stock standard solutions (1 /*g//iL)  - can be prepared from pure
         standard materials or purchased as certified solutions.

7.2.2.1     Prepare stock standard solutions by accurately weighing about
            0.0100 g of pure material.  Dissolve the material in toluene and
            dilute to volume in a 10 mL volumetric flask with toluene or
            acetone.  Larger volumes may be used at the convenience of the
            analyst.

7.2.2.2     When compound purity is assayed to be 97 percent or greater, the
            weight may be used without correction to calculate the
            concentration of the stock solution.  If the compound purity is
            assayed to be less than 97 percent, the weight must be corrected
            when calculating the concentration of the stock solution.  (See
            Exhibit E Analytical Standards Requirements.)

7.2.2.3     Fresh stock standards must be prepared once every six months or
            sooner, if standards have degraded or concentrated.  Stock
            standards must be checked for signs of degradation or
            concentration just prior to preparing working standards from them.

7.2.3    Secondary Dilution Standards

         Using stock standards, prepare secondary dilution standards in
         acetone that contain the compounds of interest either singly or mixed
         together.  Fresh secondary dilution standards must be prepared once
         every six months or sooner, if standards have degraded or
         concentrated.  Secondary dilution standards must be checked for signs
         of degradation or concentration just prior to preparing working
         standards from them.

7.2.4    Working Standards

7.2.4.1     Surrogate Standard Spiking Solution

            The surrogates, tetrachloro-m-xylene and decachlorobiphenyl, are
            adde'l to all standards,  samples, matrix spikes, and blanks.
            Prepare a surrogate spiking solution of 0.2 j*g/mL of each of the
            two compounds in acetone.

7.2.4.2     Matrix Spiking Solution

            Prepare a matrix spiking solution in acetone or methanol that
            contains the following pesticides at the concentrations specified:
                                   D-15/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 7
Reagents and Standards
                 Pesticide                Concentration uo/mL
                 gamma-BHC (Lindane)             0.5
                 4,4'-DDT                       1.0
                 Endrin                         1.0
                 Heptachlor                     0.5
                 Aldrin     "                   0.5
                 Dieldrin                       1.0
7.2.4.3     GPC Calibration Solution

7.2.4.3.1      Prepare a solution in methylene chloride that contains the
               following analytes at the minimum concentrations listed below:

               Analvte                     Concentration mq/mL
               Corn oil                            25.0
               Bis-2-ethylhexyl phthalate           0.5
               Methoxychlor                         0.1
               Perylene                             0.02
               Sulfur                               0.08

7.2.4.3.2      NOTE:  Sulfur is not very soluble in methylene chloride, but it
               is soluble in warm corn oil.  Therefore, one approach is to
               weigh out the corn oil, warm it,  and transfer the weighed
               amount of sulfur into the warm corn oil.  Mix it and then
               transfer into a volumetric flask with methylene chloride, along
               with the other calibration compounds.

7.2.4.4     Florisil Cartridge Check Solution

            Prepare a solution of 2,4,5-Trichlorophenol in acetone, at a
            concentration of 0.1 /*g/mL.

7.2.4.5     Resolution Check Mixture

            Prepare a mixture in hexane or iso-octane that contains the
            following pesticides and surrogates  at the concentrations listed
            below.

               Compound                  Concentration (nq/mL)
               gamma-Chlordane                10.0
               Endosulfan  I                   10.0
               4,4'-DDE                       20.0
               Dieldrin                       20.0
               Endosulfan  sulfate             20.0
               Endrin  ketone                  20.0
               Methoxychlor                   100.0
               Tetrachloro-m-xylene          20.0
               Decachlorobiphenyl             20.0
                                  D-16/PEST                             OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 7
                                                        Reagents and Standards
7.2.4.6     Performance Evaluation Mixture (PEM)

            Prepare the PEM in hexane or iso-octane at the concentration
            levels listed below.
              Compound
              gamma-BHC
              alpha-BHC
              4,4'-DDT
              beta-BHC
              Endrin
              Methoxychlor
              Tetrachloro-m-xylene
              Decachlorobiphenyl
                            Concentration (ng/mL)
                                   10.0
                                   10.0
                                  100.0
                                   10.0
                                   50.0
                                  250.0
                                   20.0
                                   20.0
7.2.4.7
Individual Standard Mixtures A and B
            The single component pesticide standards must be prepared in
            hexane or iso-octane at three concentrations for each analyte,
            including the surrogates.  Two separate calibration mixtures,  A
            and B (listed  below),  are used to ensure that each peak is
            adequately resolved.  The low point concentration corresponds  to
            the CRQL for each analyte.  The midpoint concentration must be  4
            times the low point concentration.  The high point concentration
            must be at least 16 times that of the low point, but a higher
            concentration may be chosen by the Contractor.  The high point
            concentration defines the upper end o* the concentration range  for
            which the calibration is valid.

                   Individual Standard Mixture A - Low
                   Point Concentration
                   alpha-BHC
                   Heptachlor
                   gamma-BHC
                   Endosulfan I
                   Dieldrin
                   Endrin
                   4,4' -ODD
                   4,4'-DDT
                   Methoxychlor
                   Tetrachloro-m-xylene
                   Decachlorobiphenyl
                                      5.0 ng/mL
                                      5.0 ng/mL
                                      5.0 ng/mL
                                      5.0 ng/mL
                                     10.0 ng/mL
                                     10.0 ng/mL
                                     10.0 ng/mL
                                     10.0 ng/mL
                                     50.0 ng/mL
                                      5.0 ng/mL
                                     10.0 ng/mL
                                  D-17/PEST
                                                            OLM3.0

-------
 Exhibit  D  Pesticides/Aroclors — Section  7
 Reagents and  Standards
                    Individual  Standard Mixture B - Low
                    Point  Concentration

                    beta-BHC                       5.0 ng/mL
                    delta-BHC                      5.0 ng/mL
                    Aldrin                        5.0 ng/mL
                    Heptachlor  epoxide             5.0 ng/mL
                      (exo-epoxy isomer)
                    alpha-Chlordane                5.0 ng/mL
                    gamma-Chlordane                5.0 ng/mL
                    4,4'-DDE                      10.0 ng/mL
                    Endosulfan  sulfate            10.0 ng/mL
                    Endrin aldehyde               10.0 ng/mL
                    Endrin ketone                 10.0 ng/mL
                    Endosulfan  II                 10.0 ng/mL
                    Tetrachloro-m-xylene           5.0 ng/mL
                    Decachlorobiphenyl            10.0 ng/mL

            NOTE:  Only the exo-epoxy isomer (isomer B) of heptachlor epoxide
            is used as an analytical standard.

7.2.4.8     Multicomponent Standards

            Toxaphene and Aroclor standards must be prepared individually
            except for Aroclor 1260 and Aroclor 1016 which may be combined in
            one standard mixture.   The calibration standards for the Aroclors
            must be prepared at concentrations of 100 ng/mL, except for
            Aroclor 1221 which must be prepared at 200 ng/mL.  Toxaphene must
            be prepared at 500 ng/mL.  All multicomponent standards must
            contain the surrogates at 20 ng/mL.   The Aroclor and Toxaphene
            solutions must be prepared in hexane or iso-octane.

7.2.5    Ampulated Standard Extracts

         Standard solutions purchased from a chemical supply house as
         ampulated extracts in glass vials may be retained 2 years from the
         preparation date, unless the manufacturer recommends a shorter time
         period.  Standard solutions prepared by the Contractor which are
         immediately ampulated in glass vials may be retained 2 years from the
         preparation date.  Upon breaking the glass seal, the expiration times
         listed in Sections 7.2.2 through 7.3 will apply.  The Contractor is
         responsible for assuring that the integrity of the standards have not
         degraded (see Section 7.3.5).

7.3   Storage of Standard Solutions

7.3.1    Store the stock and secondary dilution standard solutions at less
         than 4 CC but not greater than 6 °C in Teflon-lined screw cap amber
         bottles/vials. Fresh standards should be prepared every six months or
         sooner if comparison with check standards indicates a problem.

7.3.2    Store the GPC calibration solution in an amber glass bottle with a
         Teflon lined screw-cap at less than 4 °C but not greater than 6 °C

                                  D-18/PEST                            OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 7
                                                        Reagents and Standards
         and protect from light.  (Refrigeration may cause the corn oil to
         precipitate.  Before use, allow the calibration solution to stand at
         room temperature until the corn oil dissolves.)  Replace the
         calibration standard solution every six months, or more frequently if
         necessary.

7.3.3    Store all other working standard solutions in amber glass bottles or
         vials with Teflon lined screw caps at less than 4 °C but not greater
         than 6 °C and protect from light.  The standard solution must be
         checked frequently for stability.  Replace all working standard
         solutions after six months, or sooner if comparison with quality
         control check samples indicates a problem, except for the PEM
         solution which must be prepared weekly.  CAUTION:  Analysts must
         allow all standard solutions to equilibrate to room temperature
         before use.

7.3.4    Samples,  sample extracts and standards must be stored separately.

7.3.5    The Contractor is responsible for maintaining the integrity of
         standard solutions and verifying prior to use.  This means that
         standards must be brought to room temperature prior to use, checked
         for losses, and checked that all components have remained in the
         solution.
                                  D-19/PEST                             OLM3.0

-------
 Exhibit D Pesticides/Aroclors  —  Section  8
 Sample Collection,  Preservation and  Storage


 8.0    SAMPLE  COLLECTION,  PRESERVATION AND STORAGE

 8.1    Sample  Collection  and  Preservation

 8.1.1     Water  samples may be  collected in 1 L  (or  1  quart)  amber  glass
          containers,  fitted  with  screw-caps lined with  Teflon.   If amber
          containers are  not  available, the samples  should be protected from
          light.   Soil samples  may be collected  in glass containers or closed
          end  tubes  (e.g., brass sleeves)  in sufficient  quantity  to perform the
          analysis.   The  specific  requirements for site  sample collection  are
          outlined by the  Region.

 8.1.2     All  samples must be iced or refrigerated at  4  °C (±2 °C)  from the
          time of  collection  until extraction.

 8.2    Procedure for Sample Storage

 8.2.1     The  samples must be protected from light and refrigerated at 4 °C  (±2
          °C)  from the time of  receipt until 60 days after delivery of a
          complete reconciled sample data  package to the Agency.  After 60  days
          the  samples may be  disposed of in a manner that complies  with all
          applicable regulations.

 8.2.2     The  samples must be stored in an atmosphere demonstrated  to be free
          of all potential contaminants.

 8.3    Procedure for Sample Extract Storage

 8.3.1     Sample extracts must be protected from light and stored at less than
          4 °C but not greater than 6 °C  until 365 days after delivery of a
          complete reconciled data package to the Agency.

 8.3.2     Sample extracts must be stored in an atmosphere demonstrated to be
          free of  all potential contaminants.

 8.3.3     Samples,  sample extracts, and standards must be stored separately.

 8.4   Contract Required Holding Times

8.4.1    Extraction of water samples by  separatory funnel procedures must be
          completed within five days of the Validated Time of Sample Receipt
          (VTSR).  Extraction of water samples by continuous liquid-liquid
         extraction procedures must be started within five days of VTSR.
         Extraction of soil/sediment samples  by sonication must be completed
         within 10 days of VTSR.

8.4.2    As part of the Agency's QA program,  the Agency may provide
         Performance Evaluation (PE)  samples  as standard extracts which the
         Contractor is required to prepare per instructions provided by the
         Agency.  The extraction holding times (five days after VTSR for
         water,  10 days after VTSR for soil/sediment)  do not apply for PEs
         received as standard extracts.
                                  D-20/PEST                             OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 8
                                   Sample Collection, Preservation and Storage
8.4.3    Analysis of sample extracts must be completed within 40 days
         following the start of extraction.
                                  D-21/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 9
Calibration and Standardization
GC Operating Conditions/Initial Calibration
9.0   CALIBRATION AND STANDARDIZATION

9.1   Gas Chromatograph Operating Conditions

9.1.1    The following are the gas chromatographic analytical conditions.  The
         conditions are recommended unless otherwise noted.
          Carrier Gas:

          Column Flow:
          Make-up Gas:
          Injector Temperature:
          Injection:
          Injection Volume:
          Injector:
          Initial Temperature:
          Initial Hold Time:
          Temperature Ramp:
          Final Temperature:
          Final Hold Time:
                           Helium  (hydrogen may be used, see
                           6.23.3.7)
                           5 mL/min
                           P-5/P-10 or NZ (required)
                           > 200 °C (see Section 9.1.4)
                           On-column
                           1 or 2 jttL (see Section 9.1.3)
                           Grob-type,  splitless
                           150 °C
                           % min
                           5 C° to 6 C°/min
                           275 °C
                           Until after decachlorobiphenyl has
                           eluted  (approximately 10 minutes)
9.1.2    Optimize GC conditions for analyte separation and sensitivity.  Once
         optimized, the same GC conditions must be used for the analysis of
         all standards, samples, blanks and MS/MSD.  The linearity of the ECD
         may be greatly dependent on the flow rcte of the make-up gas.  Care
         must be taken to maintain stable and appropriate flow of make-up gas
         to the detector.

9.1.3    Manual injections must be 2 /^L.  Auto injectors may use 1 ftL volumes.
         The same injection volume must be used for all standards, blanks, and
         samples, including MS/MSD.

9.1.4    Cold (ambient temperature) on-column injectors that allow injection
         directly onto a 0.53 mm ID column may be used as long as the
         acceptance criteria for resolution, calibration, and analyte
         breakdown are met.
9.2
Initial Calibration
9.2.1    Summary of Initial Calibration

         Prior to the analysis of samples, including MS/MSD and required
         blanks, each GC/EC system must be initially calibrated at a minimum
         of three concentrations to determine instrument sensitivity and the
         linearity of response utilizing single component target compound and
         surrogate standards.  Multicomponent target compounds are calibrated
         at a single point.
                                   D-22/PEST
                                                                  OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
 9.2.2     Frequency of  Initial Calibration

          Each GC/ECD system must be  initially calibrated upon award of the
          contract, whenever major  instrument maintenance or modification is
          performed (e.g., column replacement or repair, cleaning or
          replacement of ECD, etc.) or  if the calibration verification
          technical acceptance criteria have not been met.

 9.2.3     Procedure for Initial Calibration

 9.2.3.1     Set up the GC/ECD systems  as described in Section 9.1

 9.2.3.2     Prepare the initial calibration standards using the analytes and
            the concentrations specified in Section 7.2.4.5 through 7.2.4.8.

 9.2.3.3     All standards, samples,  MS/MSD, blanks and extracts must be
            allowed to warm to ambient temperature before analysis.

 9.2.3.4     Analyze the initial calibration sequence as given below.  NOTE:
            Steps 16 and 17 are used as part of the calibration verification
            as well (see Section 9.3).

                   INITIAL  CALIBRATION SEQUENCE
                   1.         Resolution Check
                   2.         Performance Evaluation Mixture
                   3.         Aroclor  1016/1260
                   4.         Aroclor  1221
                   5.         Aroclor  1232
                   6.         Aroclor  1242
                   7.         Aroclor  1248
                   8.         Aroclor  1254
                   9.         Toxaphene
                   10.        Low  Point Standard A
                   11.        Low  Point Standard B
                   12 .        Midpoi r.t Standard A
                   13.        Midpoint Standard B
                   14.        High Point Standard A
                   15.        High Point Standard B
                   16.        Instrument Blank
                   17.        Performance  Evaluation Mixture

9.2.4    Calculations for Initial Calibration

9.2.4.1     During the initial calibration sequence,  absolute retention times
            (RT) are determined for all single component pesticides, the
            surrogates, and at least three major peaks of each multicomponent
            analyte.

9.2.4.2     For single component pesticides,  an RT is measured in each of
            three calibration standards and the mean RT is calculated as the
            average of the three values.  An RT is measured for the surrogates


                                   D-23/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 9
Calibration and Standardization
Initial Calibration
            in each of the three analyses of Individual Standard Mixture A
            during the initial calibration and the mean RT is calculated as
            the average of the three values.  Calculate a mean absolute

            retention time  (ET)  for each  single component pesticide  and
            surrogate using Equation 1.

            EQ. 1
                                        RT =
                                               n
            Where,


            ET    =     Mean absolute retention time of analyte.
            RTj    =     Absolute retention time of analyte.
            n     =     Number of measurements (3).
9.2.4.3     A retention time window is calculated for each single component
            analyte and surrogate and for the major peaks (3 to 5) of each
            multicomponent analyte by using the list in Table 1.  Windows are
            centered around the mean absolute retentior time for the analyte
            est olished during the initial calibrations.  Analytes are
            identified when peaks are observed in the RT window for the
            compound on both GC columns.

9.2.4.4     The linearity of the instrument is determined by calculating a
            percent relative standard deviation (%RSD) of the calibration
            factors from a three-point calibration curve for each single
            component pesticide and surrogate.  Either peak area or peak
            height may be used to calculate calibration factors used in the
            %RSD equation.  For example,  it is permitted to calculate
            linearity for endrin based on peak area and to calculate linearity
            for aldrin based on peak height.  It is not permitted within a
            %RSD calculation for an analyte to use calibration factors
            calculated from both peak area and peak height.  For example, it
            is not permitted to calculate the calibration factor for the low
            point standard for endrin using peak height and calculate the
            midpoint and high point standard calibration factors for endrin
            using peak area.

9.2.4.5     Calculate the calibration factor for each single component
            pesticide and surrogate over the initial calibration range using
            Equation 2.  The calibration factors for the surrogates are
            calculated from the three analyses of Individual Standard Mixture
            A only.
                                  D-24/PEST                             OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
9.2.4.6     Calculate the mean and the %RSD of the calibration factors for
            each single component pesticide and surrogate over the initial
            calibration range using Equations 3 and 4.
            EQ.  2
                         CF _  Peak area (or height] of the standard
                                       Mass injected  (ng)
            EQ.  3
                                        CF =
            EQ. 4
                                            SD
                                     %RSD = 	—  x 100
                                             CF
            Where,
            SD,.
                    \
                            2 - CF)1
             %RSD =    Percent  relative  standard  deviation
             SDCF =    Standard deviation  of  calibration  factors
             CFj   =   Calibration  factor

             CF   =    Mean  calibration  factor
             n    =    Total number of values (3)
9.2.4.7     A calibration factor is calculated for each peak in a selected set
            of three to five major peaks for each multicomponent analyte using
            Equation 2.

9.2.4.8     Calculate the percent breakdown of DDT, the percent breakdown of
            endrin and the combined breakdown of DDT and endrin in the PEM
            using Equations 5, 6, 7 and 8.
                                   D-25/PEST
OLM3.0

-------
 Exhibit D  Pesticides/Aroclors  —  Section  9
 Calibration  and  Standardization
 Initial Calibration
             EQ.  5

            Amount found (ng)  =  Peak area (hei^h^  of compound in PEM
            Where,
                 =  The calibration factor for the compound determined from
                    the midpoint standard in the most recent initial
                    calibration.  NOTE:  If during the initial calibration,
                    linearity was determined based on peak area for the
                    compound, then the midpoint CF must be based on peak area.
                    If during the initial calibration, the linearity for the
                    compound was determined based on peak height for the
                    compound, then the midpoint CF must be based on peak
                    height.
            EQ. 6
               ^Breakdown DDT =  **"Q""t found (ng) (DDD+DDE)
                                Amount (ng)  of DDT injected
            EQ. 7

  ^Breakdown Endrin =  A"70"^ found (ng) (endrin aldehyde + endrin ketone)
                                Amount (ng) of endrin injected
            EQ. 8

            Combined ^Breakdown = ^Breakdown DDT +• ^Breakdown Endrin



9.2.4.9     Calculate the percent difference for each single component
            pesticide and surrogate in the PEM using Equations 5 and 9.

            EQ. 9

                                %£> = C°alc~ C™m  x 100
            Where,

            %D   =  Percent  difference
            cnom =   Nominal  concentration of  each analyte
            ccalc=   Calculated  concentration  of  each analyte from the analyses
                    of  the standards.
                                   D-26/PEST                            OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
9.2.4.10    Calculate the resolution between the analytes in the Resolution
            Check Mixture, the Performance Evaluation Mixture, and the
            midpoint concentration of Individual Standard Mixtures A and B
            using Equation 10.
                       AIM (or Hdflht)
T
            EQ. 10
                                   ^Resolution = — x 100
                                                 H
            Where,

            V =   Depth of the valley between the two peaks. The depth of the
                  valley is measured along a vertical line from the level of
                  the apex of the shorter peak to the floor of the valley
                  between the two peaks.
            H =   Height of the shorter of the adjacent peaks.
9.2.5    Technical Acceptance Criteria for Initial Calibration

         All initial calibration technical acceptance criteria apply
         independently to both GC columns.

9.2.5.1     The initial calibration sequence must be analyzed according to the
            procedure and in the order listed in Section 9.2.3, at the
            concentrations listed in Sections 7.2.4.5 through 7.2.4.8, and at
            the frequency listed in Section 9.2.2.  The GC/ECD operating
            conditions optimized in Section 9.1 must be followed.

9.2.5.2     The resolution between two adjacent peaks in the Resolution Check
            Mixture must be greater than or equal to 60.0 percent.

9.2.5.3     All single component pesticide and surrogate peaks in both runs of
            the PEM must be greater than or equal to 90.0 percent resolved on
            each column.
                                   D-27/PEST
                             OLM3.0

-------
 Exhibit  D  Pesticides/Aroclors — Section 9
 Calibration  and  Standardization
 Initial  Calibration
 9.2.5.4     The absolute retention times of each of the single component
            pesticides and surrogates in both runs of the PEM must be within
            the retention time windows determined from the three-point initial
            calibration in Section 9.2.4.3.

 9.2.5.5     The percent difference of the calculated amount (amount found) and
            the nominal amount (amount added) for each of the single component
            pesticides and surrogates in both of the PEM runs of each GC
            column must be greater than or equal to -25.0 AND less than or
            equal to 25.0 percent using Equation 9.

 9.2.5.6     The percent breakdown of DDT and endrin in each of the PEM runs
            must be less than or equal to 20.0 percent.  The combined
            breakdown of DDT and endrin must be less than or equal to 30.0
            percent.

 9.2.5.7     The %RSD of the calibration factors for each single component
            target compound must be less than or equal to 20.0 percent,  except
            alpha-BHC and delta-BHC.   The %RSD of the calibration factors for
            alpha-BHC and delta-BHC must be less than or equal to 25.0
            percent.  The %RSD of the calibration factors for the two
            surrogates must be less than or equal to 30.0 percent.  Up to two
            single component target compounds (but not surrogates) per column
            may exceed the 20.0 percent limit for %RSD (25.0 percent for
            alpha-BHC and delta-BHC),  but those compounds must have a %RSD of
            less than or equal to 30.0 percent.

9.2.5.8     The resolution between any two adjacent peaks in the midpoint
            concentrations of Individual Standard Mixtures A and B in the
            initial calibration must  be greater than or equal  to 90.0 percent.

9.2.5.9     All instrument blanks must meet the technical acceptance criteria
            in Section 12.1.4.4.

9.2.5.10    The identification of single component pesticides  by gas
            chromatographic method^ is based primarily on retention time data.
            The retention time of the apex of a peak can only  be verified from
            an on-scale chromatogram.   The identification of muIticomponent
            analytes by gas chromatographic methods is based primarily on
            recognition of patterns of retention times displayed on a
            chromatogram.   Therefore,  the following requirements apply to all
            data presented for single component and multicomponent analytes.

            •   The chromatograms that result from the analyses of the
                Resolution Check Mixture,  the PEM,  and Individual Standard
                Mixtures  A and B during the initial calibration sequence must
                display the single component analytes present  in each standard
                at greater than 10 percent of full scale but less than 100
                percent of full scale.

            •   The chromatograms for at least one of the three analyses each
                of Individual Standard Mixtures A and B from the initial
                calibration sequence  must display the single component

                                  D-28/PEST                             OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 9
                                               Calibration and Standardization
                                                           Initial Calibration
                analytes at greater than 50 percent and less than 100 percent
                of full scale.

            •   The chromatograms of the standards for the multicomponent
                analytes analyzed during the initial calibration sequence must
                display the peaks chosen for identification of each analyte at
                greater than 25 percent and less than 100 percent of full
                scale.

            •   For any standard containing alpha-BHC, the baseline of the
                chromatogram must return to below 50 percent of full scale
                before the elution time of alpha-BHC, and return to below 25
                percent of full scale after the elution time of alpha-BHC and
                before the elution time of decachlorobiphenyl.

            •   If a chromatogram is replotted electronically to meet
                requirements, the scaling factor  
-------
Exhibit. D Pesticides/Aroclors — Section 9
Calibration and Standardization
Calibration Verification
            initial calibration technical acceptance criteria have been met
            will require reanalysis at no additional cost to the Agency.

9.3   Calibration Verification

9.3.1    Summary of Calibration Verification

         Three types of analyses are used to verify the calibration and
         evaluate instrument performance.  The analyses of instrument blanks,
         PEMs, and the midpoint concentration of Individual Standard Mixtures
         A and B constitute the continuing calibration.  Sample data are not
         acceptable unless bracketed by acceptable analyses of instrument
         blanks, PEMs, and both Individual Standard Mixtures A and B.

9.3.2    Frequency of Calibration Verification

9.3.2.1     An instrument blank and the PEM must bracket one end of a 12-hour
            period during which sample data are collected, and a second
            instrument blank and the midpoint concentration of Individual
            Standard Mixtures A and B must bracket the other end of the 12-
            hour period.

9.3.2.2     For the 12-hour period immediately following the initial
            calibration sequence, the instrument blank and the PEM that are
            the last two steps in the initial calibration sequence bracket the
            front end of that 12-hour period.  The injection of the instrument
            blank starts the beginning of that 12-hour period (see Section
            10.2.2.1).  Samples may be injected for 12 hours from the
            injection of the instrument blank.  The three injections
            immediately after that 12-hour period must be an instrument blank,
            Individual Standard Mixture A, and Individual Standard Mixture B.
            The instrument blank must be analyzed first, before either
            standard.  The Individual Standard Mixtures may be analyzed in
            either order (A,B or B,A).

9.3.2.3     The analyses of the instrument blank and Individual Standard
            Mixtures A and B immediately following one 12-hour period may be
            used to begin the subsequent 12-hour period, provided that they
            meet the acceptance criteria in Section 9.3.5.  In th?°t instance,
            the subsequent 12-hour period must be bracketed by the acceptable
            analyses of an instrument blank and a PEM, in that order.  Those
            two analyses may in turn be used to bracket the front end of yet
            another 12-hour period.  This progression may continue every 12
            hours until such time as any of the instrument blanks, PEMs, or
            Individual Standard Mixtures fails to meet the acceptance criteria
            in Sections 9.3.5.  The 12-hour time period begins with the
            injection of the instrument blank.

9.3.2.3.1      Standards (PEM or Individual Standard Mixtures), samples and
               required blanks may be injected for 12 hours from the time of
               injection of the instrument blank.
                                   D-30/PEST                             OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 9
                                               Calibration and Standardization
                                                      Calibration Verification
9.3.2.4     If more than 12 hours have elapsed since the injection of the
            instrument blank that bracketed a previous 12-hour period, an
            acceptable instrument blank and PEM must be analyzed in order to
            start a new sequence.  This requirement applies even if no
            analyses were performed since that standard was injected.

9.3.2.5     The requirements for running the instrument blanks, PEM, and
            Individual Standard Mixtures A and B are waived when no samples,
            dilutions, reanalyses, method/sulfur blanks, MS/MSD, and
            multicomponent analytes for the 72-hour confirmation requirement
            are analyzed during that 12-hour period.  To resume analysis,
            using the existing initial calibration, the Contractor first must
            analyze an instrument blank and PEM which meet the technical
            acceptance criteria.

9.3.2.6     If the entire 12-hour period is not required for the analyses of
            all samples to be reported, the sequence must be ended with either
            the instrument blank/PEM combination or the instrument
            blank/Individual Standard Mixtures A and B combination, whichever
            was due to be performed at the end of the 12-hour period.

9.3.3    Procedure for Calibration Verification

9.3.3.1     Analyze the PEM,  instrument blank and the midpoint concentration
            of Individual Standard Mixtures A and B at the required
            frequencies (Section 9.3.2).

9.3.3.2     All standards and blanks must be at ambient temperature at the
            time of preparation and analysis.

9.3.4    Calculations for Calibration Verification

9.3.4.1     For each analysis of the PEM used to demonstrate continuing
            calibration,  calculate the percent difference between the amount
            of each analyte (including the surrogates)  found in the PEM and
            the nominal amount using Equations 5 and 9.

9.3.4.2     Calculate the percent breakdown of DDT and endrin, and the
            combined breakdown in each PEM analyzed using Equations 5, 6, 7,
            and 8.

9.3.4.3     For each analysis of the midpoint concentration of Individual
            Standard Mixtures A and B used to demonstrate continuing
            calibration,  calculate the percent difference between the amount
            of each analyte (including the surrogates)  found in the standard
            mixture and the nominal amount using Equations 5 and 9.  Do not
            attempt to calculate the breakdown of Endrin and DDT in the
            Individual Standard Mixtures,  as these standards contain the
            breakdown products as well as the parent compounds.
                                  D-31/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 9
Calibration and Standardization
Calibration Verification
9.3.5    Technical Acceptance Criteria for Calibration Verification

         All calibration verification technical acceptance criteria apply
         independently to each column.  Each column roust meet criteria.

9.3.5.1     The PEMs, Individual Standard Mixtures A and B and instrument
            blanks must be analyzed at the required frequency on a GC/EC
            system that has met the initial calibration technical acceptance
            criteria.

9.3.5.2     All single component pesticides and surrogates in the PEMs used to
            demonstrate continuing calibration must be greater than or equal
            to 90.0 percent resolved.  The resolution between any two adjacent
            peaks in the midpoint concentrations of Individual Standard
            Mixtures A and B in the initial calibration must be greater than
            or equal to 90.0 percent.

9.3.5.3     The absolute retention time for each of the single component
            pesticides and surrogates in the PEMs and midpoint concentration
            of the Individual Standard Mixtures used to demonstrate continuing
            calibration must be within the retention time window determined
            from the three-point initial calibration in Section 9.2.4.3.

9.3.5.4     The percent difference between the calculated amount and the
            nominal amount (amount added)  for each of the single component
            pesticides and surrogates in the PEM must be greater than or equal
            to -25.0 percent and less than or equal to 25.0 percent.

9.3.5.5     The percent difference between the calculated amount and the
            nominal amount (amount added)  for each of the single component
            pesticides and surrogates in the INDA and INDB that have been used
            as calibration verification must be greater than or equal to -25.0
            percent and less than or equal to 25.0 percent.

9.3.5.6     The percent breakdown of DDT in the PEM must be less than or equal
            to 20.0 percent on e^ch column.  The percent breakdown of endrin
            in the PEM must be less than or equal to 20.0 percent on each
            column.  The combined breakdown of both DDT and endrin must be
            less than or equal to 30.0 percent on each column.

9.3.5.7     All instrument blanks must meet the technical acceptance criteria
            in Section 12.1.4.4.

9.3.5.8     The identification of single component pesticides by gas
            chromatographic methods is based primarily on retention time data.
            Since the retention time of the apex of a. peak can only be
            verified from an on-scale chromatogram, the following requirements
            must be met for continuing calibration to be acceptable.

9.3.5.8.1      The chromatograms that result from the analyses of the PEM and
               the Individual Standard Mixtures must display the single
               component analytes present in each standard at greater than 10
               percent of full scale but less than 100 percent of full scale.

                                   D-32/PEST                            OLM3.0

-------
                                    Exhibit D Pesticides/Aroclors — Section 9
                                               Calibration and Standardization
                                                      Calibration Verification
9.3.5.8.2      For any PEM, Individual Standard Mixture or blank, the baseline
               of the chromatogram must return to below 50 percent of full
               scale before the elution time of alpha-BHC, and return to below
               25 percent of full scale after the elution time of alpha-BHC
               and before the elution time of deca-chlorobiphenyl.

9.3.5.8.3      If a chromatogram is replotted electronically to meet these
               requirements, the scaling factor used must be displayed on the
               chromatogram.

9.3.5.8.4      If the chromatogram of any standard or blank needs to be
               replotted electronically to meet these requirements, both the
               initial chromatogram and the replotted chromatogram(s) must be
               submitted in the data package.

9.3.6    Corrective Action for Calibration Verification

9.3.6.1     If the technical acceptance criteria for the calibration
            verification are not met, inspect the system for problems and take
            corrective action to achieve the acceptance criteria.

9.3.6.2     Major corrective actions such as replacing the GC column or baking
            out the detector will require that a new initial calibration be
            performed that meets the technical acceptance criteria in Section
            9.2.5.

9.3.6.3     Minor corrective actions may not require performing a new initial
            calibration, provided that a new analysis of the standard (PEM or
            Individual Standard Mixture) that originally failed the criteria
            and an associated instrument blank immediately after the
            corrective action do meet all the acceptance criteria.

9.3.6.4     If a PEM or Individual Standard Mixture does not meet technical
            acceptance criteria listed above, it must be reinjected
            immediately.  If the second injection of the PEM or Individual
            Standard Mixture meets the criteria, sample analysis may continue.
            If the second injection does not meet the criteria, all data
            collection must be stopped.   Appropriate corrective action must be
            taken, and a new initial calibration sequence must be run before
            more sample data are collected.

9.3.6.5     If an instrument blank does not meet the technical acceptance
            criteria listed in Section 12.1.4.4, all data collection must be
            stopped.   Appropriate corrective action must be taken to clean out
            the system,  and an acceptable instrument blank must be analyzed
            before more sample data are collected.

9.3.6.6     Analysts are reminded that running an instrument blank and a PEM
            or Individual Standard Mixtures once every 12 hours is the minimum
            contract requirement.   Late eluting peaks may carry over from one
            injection to the next if highly complex samples are analyzed or if
            the GC conditions are unstable.  Such carryover is unacceptable.
                                   D-33/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 9
Calibration and Standardization
Calibration Verification
            Therefore, it may be necessary to run instrument blanks and
            standards more often to avoid discarding data.

9.3.6.7     If a successful instrument blank and PEM cannot be run after an
            interruption in analysis (Section 9.3.2.5),  an acceptable initial
            calibration must be run before sample data may be collected.  All
            acceptable sample analyses must be preceded and followed by
            acceptable standards and instrument blanks,  as described in
            Section 9.3.2.

9.3.6.8     Calibration verification technical acceptance criteria must be met
            before any samples, including MS/MSD and required blanks are
            reported.  Any samples, including KS/MSD and required blanks
            associated with a calibration verification which did not meet the
            technical acceptance criteria will require reanalysis at no
            additional cost to the Agency.
                                  D-34/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Sample Preparation
10.0  PROCEDURE

10.1  Sample Preparation

10.1.1   If insufficient sample amount  (less than 90% of the required amount)
         is received to perform the analyses, the Contractor shall contact SMO
         to apprise them of the problem.  SMO will contact the Region for
         instructions.  The Region will either require that no sample analyses
         be performed or will require that a reduced volume be used for the
         sample analysis.  No other changes in the analyses will be permitted.
         The Contractor shall document the Region's decision in the SDG
         Narrative.

10.1.2   If multiphase samples (e.g., a two-phase liquid sample, oily
         sludge/sandy soil sample) are received by the Contractor, the
         Contractor shall contact SMO to apprise them of the type of sample
         received.  SMO will contact the Region.  If all phases of the sample
         are amenable to analysis, the Region may require the Contractor to do
         the following:

         •   Mix the sample and analyze an aliquot from the homogenized
             sample.

         •   Separate the phases of the sample and analyze each phase
             separately.  SMO will provide EPA sample numbers for the
             additional phases.

         •   Separate the phases and analyze one or more of the phases, but
             not all of the phases.  SMO will provide EPA sample numbers for
             the additional phases, if required.

         •   Do not analyze the sample.

10.1.2.1    If all of the phases are not amenable to analysis (i.e.,  outside
            scope), the Region may require the Contractor to do the following:

            •   Separate the phases and analyze the phase(s) that is amenable
                to analysis.   SMO will provide EPA sample numbers for the
                additional phases, if required.

            •   Do not analyze the sample.

10.1.2.2    No other change in the analyses will be permitted.   The Contractor
            shall document the Region's decision in the SDG Narrative.

10.1.3   Extraction of Water Samples

         Water samples may be extracted by either a separatory funnel
         procedure or a continuous liquid-liquid extraction procedure.  If an
         emulsion prevents acceptable solvent recovery with the separatory
         funnel procedure,  continuous liquid-liquid extraction must be
         employed.
                                  D-35/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Sample Preparation
10.1.3.1    Separatory Funnel Extraction

10.1.3.1.1     Measure out each 1 L sample aliquot in a separate graduated
               cylinder.  Measure and record pH of the sample with wide range
               pH paper and adjust the pH to between 5 and 9 with 10 N sodium
               hydroxide or concentrated sulfuric acid, if required.  Samples
               requiring pH adjustment must be noted in the SDG Narrative.
               Place the sample into a 2 L separatory funnel.

10.1.3.1.2     Using a syringe or a volumetric pipet, add 1.0 mL of the
               surrogate solution to all water samples.

10.1.3.1.3     Rinse the graduated cylinder with 30 mL of methylene chloride
               and transfer the rinsate to the separatory funnel.  If the
               sample was received in a 1 L container, rinse the empty
               container with 30 mL of methylene chloride and add rinsate to
               the separatory funnel.  If the sample container is not rinsed,
               then add another 30 mL of methylene chloride to the separatory
               funnel and extract the sample by shaking the funnel for two
               minutes, with periodic venting to release excess pressure.
               NOTE: The total volume of solvent used for extraction is 60 mL.
               Allow the organic layer to separate from the water phase for a
               minimum of 10 minutes.  If the emulsion interface between
               layers is more than one-third the volume of the solvent layer,
               the analyst must employ mechanical techniques to complete the
               phase separation.  The optimum technique depends upon the
               sample,  and may include stirring,  filtration of the emulsion
               through glass wool, centrifugation or other physical means.
               Drain the methylene chloride into a 250 mL Erlenmeyer flask.

10.1.3.1.4     Add a second 60 mL volume of methylene chloride to the
               separatory funnel and repeat the extraction procedure a second
               time, combining the extracts in the Erlenmeyer flask.  Perform
               a third extraction in the same manner.

10.1.3.2    Continuous Liquid-Liquid Extraction

10.1.3.2.1     Continuous Liquid-Liquid Extraction Without Hydrophobic
               Membrane

10.1.3.2.1.1      Follow manufacturer's instructions for set-up.

10.1.3.2.1.2      Add methylene chloride to the bottom of the extractor and
                  fill  it to a depth of at least one inch above the bottom
                  sidearm.

10.1.3.2.1.3      Measure out each 1 L sample aliquot in a separate, clean
                  graduated cylinder; transfer the aliquot to the continuous
                  extractor.   Measure the pH of the sample with wide range pH
                  paper and record pH.   Adjust the pH to between 5 and 9 with
                  10 N sodium hydroxide or concentrated sulfuric acid, as
                  required.  Samples requiring the pH adjustment must be noted
                  in the SDG Narrative.  NOTE: With some samples, it may be

                                  D-36/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Sample Preparation
                  necessary to place a layer of glass wool between the
                  methylene chloride and the water layer in the extractor to
                  prevent precipitation of suspended solids into the methylene
                  chloride during extraction.

10.1.3.2.1.4      Using a syringe or volumetric pipet, add 1.0 mL of the
                  surrogate standard spiking solution (7.2.4.1) into the
                  sample and mix well.

10.1.3.2.1.5      Rinse the graduated cylinder with 50 mL of methylene
                  chloride and transfer the rinsate to the continuous
                  extractor.  If the sample was received in a 1 L container,
                  rinse the empty container with 50 mL of methylene chloride
                  and add the rinsate to the continuous extractor.

10.1.3.2.1.6      Add sufficient methylene chloride to the continuous
                  extractor to ensure proper solvent cycling during operation.
                  Adjust the drip rate to 5 to 15 mL/minute (recommended);
                  optimize the extraction drip rate.  Extract for a minimum of
                  18 hours.  NOTE:  When a minimum drip rate of 10-15
                  mLs/minute is maintained throughout the extraction, the
                  extraction time may be reduced to a minimum of 12 hours.
                  Allow to cool, then detach the distillation flask.  Proceed
                  to Section 10.1.4.

10.1.3.? 1.7      NOTE:  Some continuous liquid-liquid extractors are also
                  capable of concentrating the oxtract within the extraction
                  set-up.  Follow the manufacturer's instructions for
                  concentration when using this type of extractor.

10.1.3.2.2     Continuous Liquid-Liquid Extraction With Hydrophobic Membrane

10.1.3.2.2.1      Follow the manufacturer's instructions for set-up.

10.1.3.2.2.2      Measure out each 1 L sample aliquot in a separate, clean
                  graduated cylinder.  If the sample was received in a 1 L
                  container, rinse the empty container with 50 mL of methylene
                  chloride and add the rinsate to the continuous extractor.
                  If the sample was not received in a 1 L container, add 50 mL
                  of methylene chloride to the continuous extractor.  Slowly
                  transfer the aliquot to the continuous extractor.  Measure
                  the pH of the sample with wide range pH paper and record pH.
                  Adjust the pH to between 5 and 9 with 10 N sodium hydroxide
                  or concentrated sulfuric acid, as required.   Samples
                  requiring the pH adjustment must be noted in the SDG
                  Narrative.

10.1.3.2.2.3      Using a syringe or volumetric pipet, add 1.0 mL of the
                  surrogate standard spiking solution (7.2.4.1) into the
                  sample and mix well.
                                   D-37/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Sample Preparation
10.1.3.2.2.4      Rinse the graduated cylinder with 50 mL of methylene
                  chloride and transfer the rinsate to the continuous
                  extractor.

10.1.3.2.2.5      Add sufficient methylene chloride to the continuous
                  extractor to ensure proper solvent cycling during operation.
                  Adjust the drip rate to 15 mL/minute (recommended); optimize
                  the extraction drip rate.  Extract for a minimum of 6 hours.
                  (NOTE: Due to the smaller volume of solvent used during the
                  extraction process, some sample matrices (e.g.,  oily
                  samples, samples containing a high concentration of
                  surfactants) may create an emulsion which will  consume the
                  solvent volume, preventing the efficient extraction of the
                  sample.  When this occurs, add additional solvent to assure
                  efficient extraction of the sample,  and extend  the
                  extraction time for a minimum of 6 hours.   If the sample
                  matrix prevents the free flow of solvent through the
                  membrane, then the non-hydrophobic membrane continuous
                  liquid-liquid type extractor must be used.)  Allow to cool,
                  then detach the distillation flask.

10.1.3.2.2.6      NOTE:  Some continuous liquid-liquid extractors are also
                  capable of concentrating the extract within the extraction
                  set-up.  Follow the manufacturer's instructions for
                  concentration when using this type of extractor.  Using the
                  hydrophobic membrane,  it may not be necessary to dry the
                  extract with sodium sulfate.

10.1.3.2.3     NOTE:   If low surrogate recoveries occur,  assure  1) the
               apparatus was properly assembled to prevent leaks;  2) the drip
               rate/solvent cycling was optimized; and 3)  there was proper
               cooling for condensation of solvent.

10.1.3.2.4     NOTE:   Alternate continuous liquid-liquid extractor types that
               meet the requirements of the SOW may also be used.   If using
               alternate extractors or design types, follow the manufacturer's
               instructions for set-up.   Optimize the extraction  procedure.

10.1.4   Extract Drying

10.1.4.1    Assemble a Kuderna-Danish (K-D) concentrator by attaching a 10 mL
            concentrator tube to a 500 mL evaporative flask.  Other
            concentration devices or techniques may be used in place of the
            K-D if equivalency is demonstrated for all the pesticide/Aroclor
            target compounds listed in Exhibit C.

10.1.4.2    Pour the extract through a drying column containing  about 10 cm of
            anhydrous granular sodium sulfate and collect the extract in the
            K-D concentrator.  Rinse the Erlenmeyer flask and the column with
            at least two additional 20 to 30 mL portions of methylene chloride
            to complete the quantitative transfer.
                                   D-38/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Sample Preparation
10.1.5   Soil/Sediment Samples

         Mix samples thoroughly, especially composited samples.  Discard any
         foreign objects such as sticks, leaves and rocks.  Also, decant and
         discard any standing aqueous phase.
10.1.5.1    pH Determination

10.1.5.1.1     Transfer 50 g of soil/sediment to a 100 mL beaker.  Add 50 mL
               of water and stir the solution with a magnetic stirrer for 1
               hour.  Determine the pH of the sample by using a combination
               glass electrode and pH meter while the sample is stirred.
               Report pH value on the appropriate data sheet.  If the pH of
               the soil/sediment is > 9 or < 5, document any subsequent
               problems in the analysis related to pH in the SDG Narrative,
               but do not attempt to adjust the pH of the sample.  Discard the
               portion of the sample used for pH determination.

10.1.5.1.2     NOTE: If insufficient volume of soil/sediment is received, use
               a smaller 1:1 ratio of grams of sample to mLs of water for the
               pH determination, and note in the SDG Narrative.

10.1.5.2    Percent Moisture

            Weigh 5 to 10 g of the soil/sediment to the nearest 0.01 g into a
            tared crucible or aluminum weighing pan.  Determine the weight
            percent volatilized by drying overnight at 105 °C (hereafter
            referred to as percent moisture).   After the sample is dry,  remove
            the sample and pan and allow them to cool in a desiccator before
            weighing.  Calculate the percent moisture according to Equation 11
            below.  Concentrations of individual analytes will be reported
            relative to the dry weight of soil/sediment.  CAUTION: Gases
            volatized from some soil/sediment samples may require that this
            drying procedure be carried out in a hood.
            EQ. 11

           % Moisture = fframs of wet sample - grams of dry sample x 100
                                  grams of wet sample
10.1.5.3    Soil/Sediment Extraction by Sonication

10.1.5.3.1     Tune the sonicator according to the manufacturer's directions
               prior to extracting samples by this procedure.

10.1.5.3.2     Weigh approximately 30 g of sample (to the nearest 0.1 g) into
               a 250 or 400 mL beaker and add 60 g of anhydrous sodium sulfate
               (granular).

10.1.5.3.3     Add 2.0 mL of surrogate solution to all soil/sediment samples
               by using a volumetric pipet or a syringe.  Mix the sample well.
               The sample and the added sodium sulfate should be a

                                  D-39/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Sample Preparation


               homogeneous,  granular mixture at this point.   (Twice as much of
               the surrogate solution is added to soil/sediment samples than
               to water samples.)

10.1.5.3.4     Immediately add 80 to 100 mL of 1:1 methylene chloride/acetone
               to the sample.

10.1.5.3.5     Place the bottom surface of the sonicator probe about 1/2 inch
               below the surface of the solvent but above the sediment layer.

10.1.5.3.6     Sonicate for 3 minutes using a 3/4 inch disruptor horn at full
               power (output control knob at 10) with pulse  on and percent
               duty cycle knob set at 50 percent.  Do not use a microtip.
               NOTE: These settings refer to the Model W-385.  When using a
               sonicator other then Model W-385, refer to the instructions
               provided by the manufacturer for appropriate  output settings.

10.1.5.3.7     The extracted sample can be filtered by using gravity or vacuum
               filtration.

10.1.5.3.8     For gravity filtration prepare a filtration/drying bed by
               placing a plug of glass wool in the neck of a 10 cm powder
               funnel and filling the funnel to approximately half its depth
               (4 or 5 cm) with anhydrous sodium sulfate (80-100 g) .  Decant
               the extract through the packed funnel and collect it in a
               500 mL evaporative (K-D) flask attached to a  concentrator tube.

10.1.5.3.9     For vacuum filtration, use Whatman No. 41 paper in the Buchner
               funnel.  Pre-wet the paper with methylene chloride/acetone
               before decanting the solvent.

10.1.5.3.10    Repeat the extraction two more times with additional 80 to
               100 mL portions of the 1:1 methylene chloride/acetone.  Before
               each extraction, thoroughly mix the solid residue and make
               certain that the sodium sulfate is free flowing and not a
               consolidated mass.  As required, break up large lumps with a
               clean spatula.  Decant and filter the extraction solvent after
               each sonication by using the same funnel described above.
               After the final sonication, pour the entire sample into the
               funnel and rinse the beaker  and funnel with 60 mL of 1:1
               methylene chloride/acetone.

10.1.6   Concentrating the Extract

10.1.6.1    Concentration by K-D

            Other concentration devices or techniques may be used in place  of
            the K-D if equivalency is demonstrated for all the
            pesticide/Aroclor target compounds  listed in Exhibit C.

10.1.6.1.1     Add one or two clean boiling chips to the evaporative flask  and
               attach a three-ball Snyder column.  Pre-wet the Snyder  column
               by adding about 1 mL of methylene chloride to the top of  the

                                   D-40/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                     Concentrating the Extract
               column.  Place the K-D apparatus on a hot water bath  (60 °C to
               70 °C recommended) so that the concentrator tube is partially
               immersed in the hot water and the entire lower rounded surface
               of the flask is bathed with hot vapor.  Adjust the vertical
               position of the apparatus and the water temperature as required
               to complete the concentration in 15 to 30 minutes.  At the
               proper rate of distillation, the balls of the column will
               actively chatter, but the chambers will not flood with
               condensed solvent.  When the apparent volume of liquid reaches
               3 to 5 mL for water samples (and less than 10 mL for
               soil/sediment samples), remove the K-D apparatus.  Allow it to
               drain and cool for at least 10 minutes.  DO NOT ALLOW THE
               EVAPORATOR TO GO DRY.

10.1.6.1.2     For water extracts which do not require GPC cleanup, and for
               water and soil/sediment extracts which have been through the
               GPC cleanup step, proceed with the hexane exchange described in
               Section 10.1.6.2.

10.1.6.1.3     For water extracts which require GPC cleanup, remove the Snyder
               column, rinse the flask and its lower joint, collect the
               rinsate in the concentrator tube, and adjust the volume to
               10 mL with methylene chloride.  Proceed to Section 10.1.8.1.

10.1.6.1.4     For soil/sediment extracts that have not been cleaned-up using
               GPC,  it is absolutely necessary to further reduce the volume of
               all soil/sediment extracts to 1 mL in order to remove most of
               the acetone.  This is best accomplished using the nitrogen
               evaporation technique (Section 10.1.7.2).  The presence of
               acetone will cause a dead volume to develop in the GPC column
               and thus will cause loss of surrogates and analytes during GPC
               cleanups.  Adjust the soil/sediment extract volume to 10 mL
               with methylene chloride.  Proceed to Section 10.1.8.1 for
               mandatory GPC.

10.1.6.2    Solvent Exchange into Hexane

            This procedure applies to both extracts of water samples and
            extracts of soil/sediment samples.

10.1.6.2.1     With the extract in a K-D apparatus, remove the Snyder column,
               add 50 mL of hexane and a new boiling chip, and reattach the
               Snyder column.   Pre-wet the column by adding about 1 mL of
               hexane to the top.  Concentrate the solvent extract as
               described previously (Section 10.1.6.1), but increase the
               temperature of the water bath (to between 80 and 90 °C
               recommended).  When the apparent volume of liquid reaches 3 to
               •5 mL, remove the K-D apparatus and allow it to drain and cool
               for at least 10 minutes.  DO NOT ALLOW THE EVAPORATOR TO GO
               DRY.

10.1.6.2.2     Remove the Snyder column; using 1 to 2 mL of hexane, rinse the
               flask and its lower joint into the concentrator tube.  Complete

                                  D-41/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Concentrating the Extract
               quantitative transfer of the extract to a 10 mL vial by using
               hexane.

10.1.6.2.3     For samples which have not been subjected to GPC cleanup,
               adjust the volume of che hexane extract to 10 mL.   For samples
               which have been subjected to GPC cleanup, concentrate the
               hexane extract to 5 mL using a micro Snyder column or nitrogen
               evaporation, as described in Section 10.1.7.1 or 10.1.7.2.
               Proceed to Section 10.1.8.2 for Florisil cartridge cleanup.

10.1.7   Final Concentration of Extract

         Two different techniques are permitted to concentrate the extract  to
         volume before Florisil cleanup or extract volume before  instrumental
         analysis.  They are the Micro Snyder Column and Nitrogen Evaporation
         Technique.

10.1.7.1    Micro Snyder Column Concentration

            Add another one or two clean boiling chips to the concentrator
            tube and attach a two-ball micro Snyder column.  Pre-wet the
            Snyder column by adding about 0.5 mL of hexane to the top of the
            column.  Place the K-D apparatus in a hot water bath  (80 °C to
            90 °C recommended) so that the concentrator tube is partially
            immersed in the hot water.  Adjust the vertical position of the
            apparatus and the water temperature as required to complete the
            con, entration in 5 to 10 minutes.  Ac the proper rate of
            distillation the balls of the column will actively chatter but  the
            chambers will not flood with condensed solvent.  When the
            apparent volume of liquid reaches 0.5 mL, remove the  K-D apparatus
            from the water bath and allow it to drain for at least 10 minutes
            while cooling.  Remove the Snyder column and rinse its flask and
            lower joint into the concentrator tube with 0.2 mL of hexane.
            Adjust the final volume with hexane to 1 or 2 mL (see Sample
            Cleanup by Florisil Cartridge, Section 10.1.8.2.3).

10.1.7.2    Nitrogen Evaporation Technique (taken from ASTM Method D 3086)

10.1.7.2.1     Place the concentrator tube in a warm water bath (30 °C to
               35 °C recommended) and evaporate the solvent volume to the
               final volume by blowing a gentle stream of clean,  dry nitrogen
               (filtered through a column of activated carbon) onto the
               solvent.  DO NOT ALLOW THE EXTRACT TO GO TO DRYNESS.  Adjust
               the soil/sediment extract volume to 10 mL with methylene
               chloride.  Proceed to Section 10.3 for mandatory GPC cleanup
               procedures.

10.1.7.2.2     Gas lines from the gas source to the evaporation apparatus must
               be stainless steel, copper, or Teflon tubing.  Plastic tubing
               must not be used between the carbon trap and the sample as it
               may introduce interferences.  The internal wall of new tubing
               must be rinsed several times with hexane and then dried prior
               to use.

                                   D-42/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Cleanup Procedures
1C.1.7.2.3     During evaporation, the tube solvent level must be kept below
               the water level of the bath.

10.1.8   Cleanup Procedures

         There are three cleanup procedures specified in this method: GPC,
         Florisil cartridge, and sulfur cleanup.  GPC must be performed for
         all soil/sediment extracts.  GPC must be performed for water extracts
         that contain higher molecular weight contaminants that interfere with
         the analysis of the target analytes.  Florisil cartridge cleanup is
         mandatory for all extracts.  Sulfur cleanup must be performed for all
         sample extracts contaminated with sulfur.  Blanks and matrix spike
         and matrix spike duplicate samples must be subjected to the same
         cleanup as the unspiked samples.

10.1.8.1    Sample Cleanup by Gel Permeation Chromatography (GPC)

10.1.8.1.1     Introduction

               Gel Permeation Chromatography (GPC) is a size exclusion cleanup
               procedure using organic solvents and hydrophobic gels in the
               separation of natural (and synthetic) macromolecules.  The
               packing gel is porous and is characterized by the range or
               uniformity (exclusion range) of that pore size.  In the choice
               of gels, the exclusion range must be larger than the molecular
               size of the molecules to be separated.

10.1.8.1.2     GPC Column Preparation

               The instructions listed below for GPC column preparation are
               for Bio Beads.  Alternate column packings may be used if 1) the
               column packings have equivalent or better performance than the
               Bio Beads and meet the technical acceptance criteria for GPC
               calibration and GPC calibration checks, and 2)  the column
               packings do not introduce contaminants/artifacts into the
               sample which interfere with the analysis of the pesticide
               compounds.  Follow the manufacturer's instructions for
               preparation of the GPC column packing.

10.1.8.1.2.1      Weigh out 70 g of Bio Beads (SX-3).  Transfer them to a 1 L
                  bottle with a Teflon-lined cap or a 500 mL separatory funnel
                  with a large bore stopcock, and add approximately 300 mL of
                  methylene chloride.  Swirl the container to ensure the
                  wetting of all beads.   Allow the beads to swell for a
                  minimum of 2 hours.  Maintain enough solvent to cover the
                  beads sufficiently at all times.  If a guard column is to be
                  used,  repeat the above with 5 g of Bio Beads in a 125 mL
                  bottle or a beaker, using 25 mL of methylene chloride.

10.1.8.1.2.2      Turn the column upside down from its normal position and
                  remove the inlet bed support plunger (the inlet plunger is
                  longer than the outlet plunger).  Position and tighten the
                  outlet bed support plunger as near the end as possible, but

                                  D-43/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
                  no closer than 5 cm (measured from the gel packing to the
                  collar).

10.1.8.1.2.3      Raise the end of the outlet tube to keep the solvent in the
                  GPC column,  or close the column outlet stopcock.   Place a
                  small amount of solvent in the column to minimize the
                  formation of air bubbles at the base of poured column
                  packing.

10.1.8.1.2.4      Swirl the bead/solvent slurry to get a homogeneous mixture
                  and,  if the  wetting was done in a quart bottle,  quickly
                  transfer  it  to a 500 mL separatory funnel with a large bore
                  stopcock.  Drain the excess methylene chloride directly into
                  the waste beaker,  and then start draining the slurry into
                  the column by placing the separatory funnel tip against the
                  column wall.  This will help to minimize bubble formation.
                  Swirl occasionally to keep the slurry homogeneous.  Drain
                  enough to fill the column.  Place the tubing from the column
                  outlet into  a waste beaker below the column,  open the
                  stopcock  (if attached), and allow the excess solvent to
                  drain. Raise the tube to stop the flow,  and close the
                  stopcock  when the top of the gel begins to look dry.  Add
                  additional methylene chloride to just rewet the gel.

10.1.8.1.2.5      Wipe  any  remaining beads and solvent from the inner walls of
                  the top of the column with a laboratory tissue.   Loosen the
                  seal  slightly on the other plunger assembly (long plunger)
                  and insert it into the column.   Make the seal just tight
                  enough so that any beads on the glass surface will be pushed
                  forward,  but loose enough so that the plunger can be pushed
                  forward.

                  CAUTION:  Do  not tighten the seal if beads are between the
                  seal  and  the glass surface because this can damage the seal
                  and cause leakage.

10.1.8.1.2.6      Compress  the column as much as possible without applying
                  excessive force.   Loosen the seal and gradually pull out the
                  plunger.   Rinse and wipe off the plunger.  Slurry any
                  remaining beads and transfer them into the column.  Repeat
                  the step  in  Section 10.1.8.1.2.5 and reinsert the plunger.
                  If the plunger cannot be inserted and pushed in without
                  allowing  beads to escape around the seal, continue
                  compression  of the beads without tightening the seal, and
                  loosen and remove the plunger as described.  Repeat this
                  procedure until the plunger is inserted successfully.

10.1.8.1.2.7      Push  the  plunger until it meets the gel,  then compress the
                  column bed about four centimeters.

10.1.8.1.2.8      Pack  the  optional 5 cm column with approximately 5 g of pre-
                  swelled beads (different guard columns may require different
                                  D-44/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Cleanup Procedures
                  amounts).  Connect the guard column to the inlet of the
                  analytical column.

10.1.8.1.2.9      Connect the column inlet to the solvent reservoir  (reservoir
                  should be placed higher than the top of the column) and
                  place the column outlet tube in a waste container.  Placing
                  a restrictor in the outlet tube will force air out of the
                  column more quickly.  A restrictor can be made from a piece
                  of capillary stainless steel tubing of 1/16" OD x 0.010" ID
                  x 2".  Pump methylene chloride through the column at a rate
                  of 5 mL/min for one hour.

10.1.8.1.2.10     After washing the column for at least one hour, connect the
                  column outlet tube, without the restrictor, to the inlet
                  side of the UV detector.  Connect the system outlet to the
                  outlet side of the UV detector.  A restrictor  (same size as
                  the one in 10.1.8.1.2.9 above) in the outlet tube from the
                  UV detector will prevent bubble formation which causes a
                  noisy UV baseline.  The restrictor will not affect flow
                  rate.  After pumping methylene chloride through the column
                  for an additional 1-2 hours, adjust the inlet bed support
                  plunger until approximately 6-10 psi backpressure is
                  achieved.  Push the plunger in to increase pressure or
                  slowly pull outward to reduce pressure.

10.1.8.1.2.11     When the GPC column is not to be used for several days,
                  connect the column outlet line to the column inlet to
                  prevent column drying and/or channeling.   If channeling
                  occurs, the gel must be removed from the column, reswelled,
                  and repoured as described above.   If drying occurs,
                  methylene chloride should be pumped through the column until
                  the observed column pressure is constant and the column
                  appears wet.   Always recalibrate after column drying has
                  occurred to verify retention volumes have not changed.
                  NOTE: The description of solvent flow rate and column
                  pr-essure applies only to the ABC GPC apparatus.
                  Laboratories using equivalent equipment must develop the
                  parameters for their apparatus which give acceptable
                  performance as described in Section 10.1.8.1.3.

10.1.8.1.3     Calibration of GPC

10.1.8.1.3.1      Summary of GPC Calibration

10.1.8.1.3.1.1       The GPC calibration procedure is based on monitoring the
                     elution of standards with a UV detector connected to the
                     GPC column.

10.1.8.1.3.1.2       The UV detector calibration procedure described in
                     10.1.8.1.3.3 is to be used for the analyses of
                     organochlorine pesticides and Aroclors listed in Exhibit
                     C.  IT MUST NOT BE USED FOR THE ANALYSIS OF GC/MS
                     EXTRACTABLES OR OTHER ANALYTES WITHOUT A RECOVERY STUDY.

                                  D-45/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
10.1.8.1.3.2      Frequency of GPC Calibration
                  Each GPC system must be initially calibrated upon award of
                  the contract, when the GPC check solution fails to meet
                  criteria, when the column is changed,  or when channeling
                  occurs.

10.1.8.1.3.3      Procedure for GPC Calibration

                  The following instructions are for the Analytical
                  Biochemical Laboratories (ABC) system.  If you are using a
                  different GPC system, consult the manufacturer's instruction
                  manual for operating instructions.  A 2 mL injection loop
                  may be used in place of a 5 mL injection loop in accordance
                  with the manufacturer's instructions.

10.1.8.1.3.3.1       Verify the flow rate by collecting column eluate for 10
                     minutes in a graduated cylinder and measure the volume,
                     which should be 45-55 mL (4.5-5.5 mL/min).   Once the flow
                     rate is within the required range,  record the column
                     pressure (should be 6-10 psi)  and room temperature.
                     Changes in pressure, solvent flow rate, and temperature
                     conditions can affect analyte retention times and must be
                     monitored.

10.1.8.1.3.3.2       Using a 10 mL syringe,  load sample loop #1 with
                     calibration solution (Section 7.2.4.3).  With the ABC
                     automated system, the 5 mL sample loop requires a minimum
                     of 8 mL of the calibration solution.  Use a firm,
                     continuous pressure to push the sample onto the loop.
                     Switch the valve so that GPC flow is through the UV flow-
                     through cell.

10.1.8.1.3.3.3       Inject the calibration solution and obtain a UV trace
                     showing a discrete peak for each component.  Adjust the
                     detector and/or recorder sensitivity to produce a UV
                     trace that meets the requirements in Section
                     10.1.8.1.3.4.  Differences between manufacturer's cell
                     volumes and detector sensitivities may require a dilution
                     of the calibration solution to achieve similar results.
                     An analytical flow-through detector cell will require a
                     much less concentrated solution than the semi-prep cell
                     and, therefore, the analytical flowthrough detector cell
                     is not acceptable for use.

10.1.8.1.3.3.4       Determine the elution times for the phthalate,
                     methoxychlor, and perylene.  Phthalate will elute first,
                     perylene, last.

10.1.8.1.3.3.5       Choose a "DUMP" time which removes > 85 percent of the
                     phthalate.  Choose a "COLLECT" time so that > 95 percent
                     of the methoxychlor is collected, and continue to collect
                                  D-46/PEST                             OLM3.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Cleanup Procedures
                     until just prior to the  elution time of sulfur.  Use a
                     "WASH" time of 10 minutes.

                     NOTE: The DUMP and COLLECT times must be adjusted to
                     compensate for the difference in volume of the lines
                     between the detector and the collection flask.

10.1.8.1.3.3.6       Reinject the calibration solution after appropriate
                     collect and dump cycles have been set, and the solvent
                     flow and column pressure have been established.

10.1.8.1.3.3.7       Measure and record the volume of collected GPC eluate in
                     a graduated cylinder.  The volume of GPC eluate collected
                     for each sample extract processed may be  used to
                     indicate problems with the system during sample
                     processing.

10.1.8.1.3.3.8       Analyze a GPC blank by loading 5 mL of methylene chloride
                     into the GPC. Concentrate the methylene chloride that
                     passes through the system during the collect cycle using
                     Kuderna-Danish (K-D) evaporator.  Exchange the solvent to
                     hexane and analyze the concentrate by GC/EC according to
                     the procedure in 10.2 (usual protocol).  Assuming that
                     the blank represents the extract from a 1 L water sample,
                     calculate the analyte concentrations using Eguation 13.

10.1.8.1.3.4      Technical Acceptance Criteria for GPC Calibration

10.1.8.1.3.4.1       The GPC system must be calibrated at the frequency
                     described in Section 10.1.8.1.3.2.  The UV trace must
                     meet the following requirements:

                     •  Peaks must be observed and should be symmetrical for
                        all compounds in the calibration solution.

                     •  Corn oil and phthalate peaks must exhibit > 85 percent
                        resolution.

                     •  Phthalate and methoxychlor peaks must exhibit > 85
                        percent resolution.

                     •  Methoxychlor and perylene peaks must exhibit > 85
                        percent resolution.

                     •  Perylene and sulfur peaks must not be saturated and
                        must exhibit > 90 percent baseline resolution.

10.1.8.1.3.4.2       The solvent flow rate and column pressure must be within
                     the ranges described in Section 10.1.8.1.3.3.1.

10.1.8.1.3.4.3       The retention times for bis  (2-ethylhexyl) phthalate and
                     perylene must not vary more than ± 5 percent between
                     calibrations.  If the retention time shift is > 5

                                   D-47/PEST                             OLM3.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
                     percent, take corrective action.  Excessive retention
                     time shifts are caused by the following:

                     •  Poor laboratory temperature control or system leaks.

                     •  An unstabilized column that requires pumping methylene
                        chloride through it for several more hours or
                        overnight.

                     •  Excessive laboratory temperatures causing outgassing
                        of the methylene chloride.

10.1.8.1.3.4.4       The analyte concentrations in a GPC blank must be less
                     than the CRQL of any compound in Exhibit C (Pesticides).

10.1.8.1.3.5      Corrective Action for GPC Calibration

10.1.8.1.3.5.1       If the flow rate and/or column pressure do not fall
                     within the ranges in Section 10.1.8.1.3.3.5,  a new column
                     should be prepared.

10.1.8.1.3.5.2       AUV trace that does not meet the criteria in Section
                     10.1.8.1.3.4 would also indicate that a new column should
                     be prepared.  It may be necessary to obtain a new lot of
                     Bio Beads if the column fails all the criteria.

10.1.8.1.3.5.3       If the GPC blank is equal to or exceeds the CRQL of any
                     compound in Exhibit C (Pesticides),  pump additional
                     methylene chloride through the system for 1-2 hours.
                     Analyze another GPC blank to ensure the system is
                     sufficiently clean.  Repeat the methylene chloride
                     pumping if necessary.

10.1.8.1.4     GPC Calibration Check

10.1.8.1.4.1      Summary of GPC Calibration Check

                  The GPC calibration must be routinely verified with two
                  check mixtures.  No Florisil cleanup is used in the GPC
                  calibration check.

10.1.8.1.4.2      Frequency of GPC Calibration Check

10.1.8.1.4.2.1       The calibration check must be performed at least once
                     every 7 days whenever samples, including matrix spikes,
                     matrix spike duplicates, and blanks are cleaned up using
                     the GPC.

10.1.8.1.4.2.2       Some samples may contaminate the SX-3 Bio Beads and
                     change the retention volume of the GPC column.
                     Therefore, system calibration and analyte recovery must
                     be checked whenever a sample causes significant
                     discoloration of the GPC column.  Even if no darkening is

                                   D-48/PEST                             OLM3.0

-------
                                    Exhibit D  Pesticides/Aroclors  —  Section  10
                                                                      Procedure
                                                             Cleanup  Procedures
                      visible,  GPC  calibration must be  checked  not  less  than
                      once  every  seven  days.  In many cases,  the  SX-3  Bio Beads
                      may be  used for several months as long  as the column
                      calibration and flow  rate remain  constant.

 10.1.8.1.4.3       Procedure  for  GPC Calibration Check

                   The instructions below are for a GPC injection loop of 5  mL.
                   If  a 2 mL  injection  loop is used, the Contractor will adjust
                   the volume to  4  mL instead of 10 mL  before injecting  extract
                   on  the GPC.

 10.1.8.1.4.3.1        The pesticide GPC calibration check solution  contains  the
                      following six compounds in methylene  chloride:   gamma-
                      BHC,  Hepatachlor, and Aldrin each at  a  concentration of
                      0.1 iig/mL for a 5 mL  GPC loop (0.25 ng/mL when a 2 mL  GPC
                      loop  is used) and 4,4 '-DDT, Endrin and  Dieldrin  at 0.2
                      jig/mL (0.25 ^g/mL for a 2 mL loop).   The  Aroclor mixture
                      contains  2 ng/mL  each of Aroclor  1016 and 1260 in
                      methylene chloride (0.25 ng/mL when a 2 mL  GPC loop is
                      used).

 10.1.8.1.4.3.2        Load  the  first 5  mL sample loop by using  a  10 mL syringe
                      containing  8  tnL of the pesticide  GPC  calibration check
                      solution.  The Aroclor mixture is  loaded  into  loop #2  in
                      the same  manner.  Fractions are collected in  an  auto
                      sequence  by using the GPC program established by the UV
                      detector  calibration  procedure (Section 10.1.8.1.3).

 10.1.8.1.4.3.3        The collected GPC calibration fraction  is transferred  to
                      a  K-D apparatus,  and the collection vessel  is rinsed with
                      two additional 10 mL portions of  methylene  chloride to
                      complete  the  transfer.  The volume of methylene  chloride
                      is reduced according to Section 10.1.6.1.   After cooling,
                      the solvent is exchanged to hexane according to  the
                      instructions  in Section 10.1.6.2.   The  final volume is
                      adjusted to 10 mL, and the sample  is  analyzed by GC
                      according to  the procedure in Section 10.2  (usual
                      protocol).  The analysis must be  performed  on only one of
                      the GC  columns used for sample analysis.

10.1.8.1.4.3.4        The pattern of the Aroclor quantitation peaks and  the
                      recovery of each single component analyte must be
                      determined for evaluation and reporting purposes.

10.1.8.1.4.4      Technical Acceptance Criteria for GPC Calibration Check

10.1.8.1.4.4.1        The GPC must meet the technical acceptance  criteria for
                      GPC calibration in Section 10.1.8.1.3.4 and be calibrated
                      at the  frequency listed in Section 10.1.8.1.4.2.

10.1.8.1.4.4.2        The recovery of each of the single component analytes
                     must be between 80 - 110 percent.

                                   D-49/PEST                             OLM03.1

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
10.1.8.1.4.4.3       The Aroclor patterns must be the same as those from the
                     Aroclor 1016 and Aroclor 1260 standards in the initial
                     calibration sequence.

10.1.8.1.4.5      Corrective Action for GPC Calibration Check

                  Analysts may continue to use the GPC column if the technical
                  acceptance criteria for the GPC calibration check are met.
                  If the recoveries are out of the acceptance window or if
                  changes in the relative peak heights of the patterns of the
                  Aroclor are observed, the columns must be replaced and the
                  GPC recalibrated according to the instructions in Section
                  10.1.8.1.3 before proceeding with any GPC cleanup on
                  samples, including blanks - method and/or sulfur and MS/MSD.

10.1.8.1.5     Daily UV Calibration Check (Optional)

               The calibration of the GPC may be monitored daily by use of the
               UV-GPC calibration solution (Section 7.2.4.3)  and the UV
               detector calibration procedure (Section 10.1.8.1.3.3).   The UV
               detector should be used to monitor the elution times for the
               phthalate, methoxychlor and perylene,  in that order.  The pre-
               calibrated GPC program should "dump" > 85 percent of the
               phthalate and should "collect" > 95 percent of the methoxychlor
               and perylene.  Significant changes in elution times of  the
               analytes (e.g.,  > 0.5 minutes) indicate that the column is out
               of calibration and must be recalibrated or replaced.

10.1.8.1.6     Sample Cleanup by GPC

10.1.8.1.6.1      Introduction to Sample Cleanup by GPC

10.1.8.1.6.1.1       It is very important to have consistent laboratory
                     temperatures during an entire GPC run, which could be 24
                     hours or more.  If temperatures are not consistent,
                     retention times will shift,  and the dump and collect
                     times determined by the calibration standard no longer
                     will be appropriate.  The ideal laboratory temperature to
                     prevent outgassing of the methylene chloride is 22 °C.

10.1.8.1.6.1.2       In order to prevent overloading of the GPC column, highly
                     viscous sample extracts must be diluted prior to cleanup.
                     Any sample extract with a viscosity greater than that of
                     1:1 glycerol:water solution must be diluted and loaded
                     into several loops.  Similarly,  extracts containing more
                     than 40 mg/mL of nonvolatile residue must be diluted and
                     loaded into several loops.  The nonvolatile residue may
                     be determined by evaporating a 100 /xL aliquot of the
                     extract to dryness in a tared aluminum weighing pan, or
                     other suitable container.  When multiple loops/runs are
                     necessary for an individual sample, be sure to combine
                     all of the sample eluates collected.
                                  D-50/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Cleanup Procedures
10.1.8.1.6.1.3       Systems using automated injection devices to load the
                     sample on the column must be carefully monitored to
                     assure that the required amount is injected onto the
                     column.  Viscous extracts, or extracts containing a large
                     amounts of nonvolatile residue, will cause problems with
                     injecting the proper amount of sample extract onto the
                     column using automated injection systems.  After the
                     sample extract has been processed, the remaining sample
                     extract in an injection vial must be checked to assure
                     that the proper amount of extract was injected on the
                     column before proceeding with the sample analysis.  If
                     the proper amount of extract was not injected, the sample
                     must be reprepared and the sample extract must be either
                     diluted and loaded into several loops or the sample
                     extract must be injected manually.

10.1.8.1.6.2      Frequency of Sample Cleanup by GPC

                  GPC cleanup must be performed at least once for each
                  soil/sediment extract, water extracts that contain high
                  molecular weight contaminants that interfere with the
                  analysis of the target analytes and all associated QC
                  (blanks and spikes).  If cleanup procedure is inadequate
                  contact SMO.

10.1.8.1.6.3      Procedure for Sample Cleanup by GPC

10.1.8.1.6.3.1       Particles greater than 5 microns may scratch the valve,
                     which may result in a system leak and cross contamination
                     of sample extracts in the sample loops.  To avoid such
                     problems, filter the extract through a 5 micron filter
                     disc by attaching a syringe filter assembly containing
                     the filter disc to a 10 mL syringe.  Draw the sample
                     extract through  the filter assembly and into the 10 mL
                     syringe.  Disconnect the filter assembly before
                     transferring the sample extract into a small glass
                     container (e.g., a 15 mL culture tube with a Teflon-lined
                     screw cap).   Alternatively, draw the extract into the
                     syringe without the filter assembly.  Attach the filter
                     assembly and force the extract through the filter and
                     into the glass container.  Draw a minimum of 8 mL of
                     extract into a 10 mL syringe.

                     Note:  Some GPC instrument manufacturer's recommend using
                     a smaller micron size filter disc.  In this instance,
                     follow the manufacturer's recommended operating
                     instructions.

10.1.8.1.6.3.2       INTRODUCTION OF PARTICULATES OR GLASS WOOL INTO THE GPC
                     SWITCHING VALVES MAY REQUIRE FACTORY REPAIR OF THE
                     APPARATUS.
                                   D-51/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
10.1.8.1.6.3.3       The following instructions are for the Analytical
                     Biochemical Laboratories (ABC) system.  If you are using
                     a different GPC system,  consult your manufacturer's
                     instruction manual for operating instructions.  Note:
                     These instructions were written for a 5 mL GPC injection
                     loop.  A 2 mL injection loop may be used in place of a 5
                     mL injection loop.  If a 2 mL loop is used, concentrate
                     the 10 mL extract to 4 mL, and then inject 2 mL from the
                     4 mL.

10.1.8.1.6.3.4       Prior to loading samples, put the GPC into the "load"
                     mode, set the instrument terminal for the number of loops
                     to be loaded, and set the "dump," "collect," and "wash"
                     times for the values determined by the calibration
                     procedure described in Section 10.1.8.1.3.

10.1.8.1.6.3.5       Using a 10 mL syringe, load ^he sample into the system.
                     With the ABC automated system, the 5 mL loop requires a
                     minimum of 8 mL of sample.  Attach the syringe to the
                     turn lock on the injection port.   Use firm, continuous
                     pressure to push the sample onto the 5 mL sample loop.
                     If the sample is difficult to load,  some part of the
                     system may be blocked.  Take appropriate corrective
                     action.   If the back pressure is normal (6-10 psi), the
                     blockage is probably in the valve.  Blockage may be
                     flushed out of the valve by reversing the inlet and
                     outlet tubes and pumping solvent through the tubes (this
                     should be done before sample loading).  NOTE:
                     Approximately 2 mL of the extract remains in the lines
                     between the injection port and the sample loop; excess
                     sample also passes through the sample loop to waste.

10.1.8.1.6.3.6       After loading a loop, and before removing the syringe
                     from the injection port, index the GPC to the next loop.
                     This will prevent loss of sample caused by unequal
                     pressure in the loops.

10.1.8.1.6.3.7       After loading each sample loop, wash the loading port
                     with methylene chloride in a PTFE wash bottle to minimize
                     cross contamination.  Inject approximately 10 mL of
                     methylene chloride to rinse the common tubes.

10.1.8.1.6.3.8       After loading all sample loops, index the GPC to the 00
                     position, switch to the "RUN" mode and start the
                     automated sequence.  Process each sample using the
                     collect and dump cycle times established in Section
                     10.1.8.1.3.

10.1.8.1.6.3.9       Collect each sample in a 250 mL Erlenmeyer flask covered
                     with aluminum foil to reduce solvent evaporation, or
                     directly into a Kuderna-Danish evaporator.  Monitor
                     sample volumes collected.  Changes in sample volumes
                                   D-52/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Cleanup Procedures
                     collected may indicate one or more of the following
                     problems:

                     •  Change in solvent flow rate, caused by channeling in
                        the column or changes in column pressure.

                     •  Increase in column operating pressure due to the
                        absorption of particles or gel fines onto either the
                        guard column or the analytical  column gel, if a guard
                        column is not used.

                     •  Leaks in the system or significant variances in room
                        temperature.

10.1.8.1.6.3.10   After the appropriate GPC fraction has been collected for
                  each sample, concentrate the extract as per Section 10.1.6.1
                  and proceed to solvent exchange into hexane as described in
                  Section 10.1.6.2 and Florisil cleanup in 10.1.8.2.

10.1.8.1.6.3.11   Any samples that were loaded into two or more loops must be
                  recombined before proceeding with concentration.

10.1.8.2    Florisil Cleanup

10.1.8.2.1     Introduction

                lorisil cartridge cleanup significantly reduces matrix
               interference caused by polar compounds and is required for all
               extracts.   The same volume of the concentrated extract taken
               for Florisil cleanup must be maintained after Florisil cleanup
               (1 or 2 mL).

10.1.8.2.2     Florisil Cartridge Performance Check

10.1.8.2.2.1      Summary of Florisil Cartridge Performance Check

                  Every lot number of Florisil cartridges must be tested
                  before they are used for sample cleanup.

10.1.8.2.2.2      Frequency of Florisil Cartridge Performance Check

                  Cartridge performance check must be conducted at least once
                  on each lot of cartridges used for sample cleanup.

10.1.8.2.2.3      Procedure for Florisil Cartridge Performance Check

                  Add 0.5 mL of 2,4,5-trichlorophenol solution (0.1 ng/mL in
                  acetone, Section 7.2.4.4) and 0.5 mL of Standard Mixture A,
                  midpoint concentration,  Section 7.2.4.7) to 4 mL of hexane.
                  Reduce the final volume to 0.5 mL using nitrogen (Section
                  10.1.7.2).  Place the mixture onto the top of a washed
                  Florisil cartridge, and elute it with 9 mL of hexane/acetone
                  [(90:10)(V/V)].  Use two additional 1 mL hexane rinses to

                                  D-53/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
                  ensure quantitative transfer of standard from the cartridge.
                  Reduce the final volume to 1 mL using nitrogen (Section
                  10.1.7.2) and analyze the solution by GC/EC using at least
                  one of the GC columns specified for sample analysis.
                  Determine the recovery of each analyte for evaluation and
                  reporting purposes.  Calculate the percent recovery using
                  Equation 12.
               EQ. 12
                                             Q.
                          Percent Recovery = —- x 100
                  Where,
                  Qd = Quantity determined by analysis
                  Qa = Quantity added
10.1.8.2.2.4      Technical Acceptance Criteria for Florisil Cartridge
                  Performance Check

10.1.8.2.2.4.1       The cartridge performance check solution must be analyzed
                     on a GC/EC meeting the initial calibration and
                     calibration verification technical acceptance criteria.

10.1.8.2.2.4.2       The lot of Florisil cartridges is acceptable if all
                     pesticides are recovered at 80 to 120 percent, if the
                     recovery of trichlorophenol is less than 5 percent, and
                     if no peaks interfering with the target analytes are
                     detected.

10.1.8.2.2.5      Corrective Action for Florisil Cartridge Performance Check

                  Any lot of Florisil cartridges that does not meet the
                  criteria above must be discarded and a new lot, meeting
                  criteria, used for sample cleanup.

10.1.8.2.3     Sample Cleanup by Florisil Cartridge

               The required Florisil cartridge size and the final volume of
               the extract after Florisil cleanup are a function of the GC
               autosampler that a laboratory uses.  If the autosampler
               operates reliably with 1 mL of sample extract, then a 500 mg
               cartridge is used and the required final volume is 1 mL.  If
               the autosampler requires more sample, prepare 2 mL of sample
               extract using a 1 g cartridge.  Manual injection requires only
               a 1 mL final extract and a 500 mg cartridge.

10.1.8.2.3.1      Frequency of Sample Cleanup by Florisil Cartridge
                                  D-54/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                            Cleanup Procedures
                  All sample extracts are required to be cleaned up by the
                  Florisil cartridge technique.

10.1.8.2.3.2      Procedure for Sample Cleanup by Florisil Cartridge

10.1.8.2.3.2.1       Attach the vacuum manifold to a water aspirator or to a
                     vacuum pump with a trap installed between the manifold
                     and the vacuum source.  Adjust the vacuum pressure in the
                     manifold to between 5 and 10 pounds of vacuum.

10.1.8.2.3.2.2       Place one Florisil cartridge into the vacuum manifold for
                     each sample extract.

10.1.8.2.3.2.3       Prior to cleanup of samples, the cartridges must be
                     washed with hexane/acetone (90:10).  This is accomplished
                     by placing the cartridge on the vacuum manifold, by
                     pulling a vacuum, and by passing at least 5 mL of the
                     hexane/acetone solution through the cartridge.  While the
                     cartridges are being washed, adjust the vacuum applied to
                     each cartridge so that the flow rate through each
                     cartridge is approximately equal.   DO NOT ALLOW THE
                     CARTRIDGES TO GO DRY AFTER THEY HAVE BEEN WASHED.

10.1.8.2.3.2.4       After the cartridges on the manifold are washed, the
                     vacuum is released,  and a rack containing labeled 10 mL
                     volumetric flasks is placed inside the manifold.  Care
                     must be taken to ensure that the solvent line from each
                     cartridge is placed inside of the appropriate volumetric
                     flask as the manifold top is replaced.

10.1.8.2.3.2.5       After the volumetric flasks are in place, the vacuum to
                     the manifold is restored,  and a volume of extract equal
                     to the required final volume (1 or 2 mL) from each
                     sample, blank or matrix spike extract is transferred to
                     the top frit of the appropriate Florisil cartridge.  This
                     must equal the final volume after Florisil cleanup.

10.1.8.2.3.2.6       Because the volumes marked on concentrator tubes are not
                     necessarily accurate at the 1 mL level, the use of a
                     syringe or a volumetric pipet is required to transfer the
                     extract to the cleanup cartridge.

10.1.8.2.3.2.7       The pesticides/Aroclors in the extract concentrates are
                     then eluted through the column with 8 mL of
                     hexane/acetone (90:10) and are collected into the 10 mL
                     volumetric flasks held in the rack inside the vacuum
                     manifold.

10.1.8.2.3.2.8       Transfer the eluate in each volumetric flask to a clean
                     centrifuge tube or 10 mL vial.  Use two additional 1 mL
                     hexane rinses to ensure quantitative transfer of the
                     cartridge eluate.
                                  D-55/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
Cleanup Procedures
10.1.8.2.3.2.9       Adjust the extract to the same 1 or 2 mL aliquot volume
                     as was taken for cleanup using either nitrogen blowdown
                     (Section 10.1.7.2) or a micro Snyder column (Section
                     10.1.7.1).  Measure the final volume with a syringe or by
                     transferring the extract to a volumetric flask.

10.1.8.2.3.2.10      If sulfur cleanup is to be performed, proceed to Section
                     10.1.8.3.  Otherwise, transfer the sample to a GC vial
                     and label the vial.  The extract is ready for GC/EC
                     analysis.

10.1.8.3    Sulfur Cleanup

10.1.8.3.1     Introduction to Sulfur Cleanup

10.1.8.3.1.1      Sulfur contamination will cause a rise in the baseline of a
                  chromatogram and may interfere with the analyses of the
                  later eluting pesticides.  If crystals of sulfur are evident
                  or if the presence of sulfur is suspected, sulfur removal
                  must be performed.  Interference which is due to sulfur is
                  not acceptable.   Sulfur can be removed by one of two
                  methods,  according to laboratory preference.  If the sulfur
                  concentration is such that crystallization occurs in the
                  concentrated extract, centrifuge the extract to settle the
                  crystals, and remove the sample extract with a disposable
                  pipette,  leaving the excess sulfur in the centrifuge tube.
                  Transfer the extract to a clear, centrifuge tube or clean
                  concentrator tube before proceeding with further sulfur
                  cleanup.

10.1.8.3.1.2      If only part of  a set of samples require sulfur cleanup,
                  then, a sulfur cleanup blank is required for that part of
                  the set (Section 12.1.3).

10.1.8.3.2     Frequency of Sulfur Cleanup

               Sulfur removal is required for all sample extracts that contain
               sulfur.

10.1.8.3.3     Procedure for Sulfur Cleanup

10.1.8.3.3.1      Mercury Technique

                  Add one to three drops of mercury to each hexane extract in
                  a clean vial.  Tighten the top on the vial and agitate the
                  sample for 30 seconds.  Filter or centrifuge the extract.
                  Pipet the extract to another vial and leave all solid
                  precipitate and  liquid mercury.  If the mercury appears
                  shiny, proceed to Section 10.2 and analyze the extract.  If
                  the mercury turns black, repeat sulfur removal as necessary.
                  The extract transferred to the vial still represents the 1.0
                  or 2.0 mL final  volume.  CAUTION: Waste  containing mercury
                  should be segregated and disposed of properly.  NOTE:

                                  D-56/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                                GC/EC Analysis
                  Mercury  is a highly toxic metal and therefore must be used
                  with great care.  Prior to using mercury, it is recommended
                  that the analyst become acquainted with proper handling and
                  cleanup techniques associated with this metal.
10.1.8.3.3.2      Copper Technique

                  Add approximately 2 g of cleaned copper powder to the
                  extract in the centrifuge or concentrator tube (2 g will
                  fill the tube to about the 0.5 mL mark).  Mix the copper and
                  extract for at least 1 minute on a mechanical shaker.
                  Separate the extract from the copper powder by drawing off
                  the extract with a disposable pipet, and transfer the
                  extract to a clean vial.  The extract transferred to the
                  vial still represents the 1.0 or 2.0 mL final volume.  The
                  separation of the extract from the copper powder is
                  necessary to prevent degradation of the pesticides.  If the
                  copper  appears bright, proceed to Section 10.2 and analyze
                  the extracts.  If the copper changes color, repeat the
                  sulfur removal procedure as necessary.

10.2  GC/EC Analysis

10.2.1   Introduction to Sample Analysis by GC/EC

10.2.1.1    Before samples or required blanks can be analyzed, the instrument
            must meet the initial calibration and calibration verification
            technical acceptance criteria.  Sample analysis on both GC columns
            is required for all samples, blanks,  matrix spikes, and matrix
            spike duplicates.

10.2.1.2    Sample extracts,  standards and blanks must be analyzed within an
            analytical sequence as defined in Section 10.2.2.1, under the same
            instrumental conditions.

10.2.1.3    Set up the GC/EC system per the requirements in Section 9.0.
            Unless ambient temperature on-column injection is used (see
            Section 9.1.4),  the injector must be heated to at least 200 °C.
            The optimized gas chromatographic conditions from Section 9.1 must
            be used,

10.2.2   Procedure for Sample Analysis by GC/EC

         The injection must be made on-column by using either automatic or
         manual injection.   If autoinjectors are used, 1 fiL injection volumes
         may be used.  Manual injections shall use at least 2 jtL injection
         volumes.  The same injection volume must be used for all standards,
         samples, MS/MSD,  and blanks  associated with the  same initial
         calibration.  If a single injection is used for two GC columns
         attached to a single injection port,  it may be necessary to use an
         injection volume greater than 2 /xL.   However, the same injection
         volume must be used for all  analyses.
                                  D-57/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
GC/EC Analysis
10.2.2.1    Analytical Sequence

            All acceptable samples must be analyzed within a valid analysis
            sequence as given below.
     Time
Injection #
Material Injected
     0 hr.
     12 hr.
     Another 12 hr.
     Another 12  hr.
1-15

16

17

18
0
0
0
0
1st injection past
12:00 hr.
2nd and 3rd
injections past
12:00 hr.
0
0
0
0
0
0
1st injection past
12 hr.
2nd injection
0
0
0
0
0
1st injection past
12:00 hr.
First 15 steps of the initial
calibration
Instrument blank at end of
initial calibration
PEM at end of initial
calibration
First sample

Subsequent samples

Last sample
Instrument blank

Individual Standard Mixtures
A and B

Sample
Subsequent samples

Last sample
Instrument blank

PEM
Sample

Subsequent samples

Last sample
Instrument blank
                                  D-58/PEST
                                                  OLM03.0

-------
                                    Exhibit  D  Pesticides/Aroclors  —  Section  10
                                                                      Procedure
                                                                GC/EC Analysis
      Time             Injection #           Material Injected

                      2nd and 3rd           Individual Standard Mixtures
                      injections past       A and B
                      12:00 hr.
                      0                     Sample
                      0
                      0
                      0                     Subsequent samples
                      0
                      etc.

10.2.2.1.1     NOTE:  The first  12 hours are counted from injection #16  (the
               instrument  blank at the end of the  initial calibration
               sequence),  not  from injection #1.   Samples may be injected
               until  12 hours  have elapsed.  All subsequent 12-hour periods
               are timed from  the injection of the instrument blank that
               brackets the  front end of the samples.  Because the 12-hour
               time period is  timed  from injection of the instrument blank
               until  the injection of the last sample, each 12-hour period may
               be separated  by  the length of one chromatographic run, that of
               the analysis  of  the last sample.  While the 12-hour period may
               not be exceeded, the  laboratory may run instrument blanks and
               standards more  frequently for instance, to accommodate staff
               working 8-hour  shifts.

10.2.2.1.2     After  the initial calibration, the analysis sequence may
               continue as long as acceptable instrument blanks, PEMs and
               Individual Standard Mixtures A and B are analyzed at the
               required frequency.  This analysis sequence shows only the
               minimum required blanks and standards.  More blanks and
               standards may be run at the discretion of the Contractor; uhese
               must also satisfy the criteria presented in Section 9 in order
               to continue the run sequence.

10.2.2.1.3     An analysis sequence must also include all required matrix
               spike/matrix  spike duplicate and method (and/or sulfur) blank
               analyses, but the Contractor may decide at what point in the
               sequence they are to be analyzed.

10.2.2.1.4     The requirements for the analysis sequence apply to both GC
               columns and for all instruments used for these analyses.

10.2.3   Sample Dilutions

10.2.3.1    All samples must be analyzed at the most concentrated level that
            is consistent with achieving satisfactory chromatography (defined
            in Section 11.3).

10.2.3.2    If the response of any single component pesticide is greater than
            the response of that analyte in the initial calibration high point

                                  D-59/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 10
Procedure
GC/EC Analysis
            standard, then the extract must be diluted to have the response of
            that analyte between the initial calibration low point and high
            point standards.

10.2.3.3    If the response of the largest peak in a multicomponent analyte is
            greater than the most intense single component analyte response in
            the initial calibration high point standard, then the sample must
            be diluted to have the response of the largest peak in the
            multicomponent analyte between the responses of the initial
            calibration midpoint and high point standards of that single
            component pesticide.

10.2.3.4    If dilution is employed solely to bring a peak within the
            calibration range or to get a multicomponent pattern on scale,  the
            results for both the more and the less concentrated extracts must
            be reported.  The resulting changes in quantitation limits and
            surrogate recovery must be reported also for the diluted samples.

10.2.3.5    If the Contractor has reason to believe that diluting the final
            extracts will be necessary, an undiluted run may not be required.
            If an acceptable chromatogram (as defined in Section 11.3) is
            achieved with the diluted extract, an additional extract 10 times
            more concentrated than the diluted sample must be injected and
            reported with the sample data.

10.2.3.6    When diluted, the chromatographic data for the single component
            pesticides must be able to be reported at greater than 10.0
            percent of full scale but less than 100.0 percent of full scale.

10.2.3.7    When diluted, multicomponent analytes must be able to be reported
            at greater than 25.0 percent of full scale but less than 100.0
            percent of full scale.

10.2.3.8    If a chromatogram is replotted electronically to meet these
            requirements, the scaling factor used must be displayed on the
            chromatogram.  If the chromatogram of any sample needs to be
            replotted electronically to meet these requirements both the
            initial chromatogram and the replotted chromatogram must be
            submitted in the data package.

10.2.3.9    Samples with analytes detected at a level greater than the high
            calibration point must be diluted until the response is within  the
            linear range established during calibration or to a maximum of
            1:100,000.

10.2.3.10   If the response is still above the high calibration point after
            the dilution of 1:100,000, the Contractor shall contact SMO
            immediately.

10.2.3.11   Use the results of the original analysis to determine the
            approximate dilution factor  required to get the largest analyte
            peak within the initial calibration range.
                                  D-60/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 10
                                                                     Procedure
                                                                GC/EC Analysis
10.2.3.12   The dilution factor chosen should keep the response of the largest
            peak for a target compound in the upper half of the initial
            calibration range of the instrument.

10.2.3.13   Sample dilutions must be made quantitatively.  Dilute the sample
            extract with hexane.

10.2.3.14   Do not submit data for more than two analyses, i.e., the original
            sample extract and one dilution, or, if a screening procedure was
            employed, from the most concentrated dilution analyzed and one
            further dilution.  This statement does not refer to reanalyses
            required due to failed technical acceptance criteria.
                                  D-61/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 11
Data Analysis and Calculations
Qualitative Identification
11.0  DATA ANALYSIS AND CALCULATIONS

11.1  Qualitative Identification

11.1.1   Identification of Target Compounds

11.1.1.1    The laboratory will identify and quantitate analyte peaks based on
            the RT windows and the calibration factors of the midpoint
            standard (single component pesticides)  established during the
            initial calibration sequence.

11.1.1.2    Analytes are identified when peaks are observed in the RT window
            for the analyte on both GC columns.

11.1.1.3    A set of three to five major peaks is selected for each
            multicomponent analyte.  The retention time window for each peak
            is determined from the initial calibration analysis.
            Identification of a multicomponent analyte in the sample is based
            on pattern recognition in conjunction with the elution of three to
            five sample peaks within the retention time windows of the
            corresponding peaks of the standard  on both GC columns.
            Calibration factors used to quantitate toxaphene and the Aroclors
            are based on the single-point calibration standard analyzed during
            the initial calibration.  The number of potential quantitation
            peaks is listed in Table 2.

11.1.1.4    When any multicomponent analyte is detected in a sample, a
            standard must be run within 72 hours of the analyte's detection
            (from time of injection),  and within a valid 12-hour sequence.

11.1.1.5    The choice of the peaks used for multicomponent analyte
            identification and the recognition of those peaks may be
            complicated by the environmental alteration of the toxaphene or
            Aroclors, and by the presence of co-eluting analytes or matrix
            interferences,  or both.  Because of  the alteration of these
            materials in the environment, multicomponent analytes in samples
            may give patterns similar to, but not identical with, those of the
            standards.

11.1.1.6    If more than one multicomponent analyte is observed in a sample,
            the Contractor must choose different peaks to quantitate each
            multicomponent analyte.  A peak common to both analytes present in
            the sample must not be used to quantitate either compound.

11.1.2   GC/MS Confirmation of Pesticides and Aroclors

11.1.2.1    Any pesticide or Aroclor analyte listed in Exhibit C for which a
            concentration is reported from a GC/EC analysis must have the
            identification confirmed by GC/MS if the concentration is
            sufficient for that purpose.  The following paragraphs are to be
            used as guidance in performing GC/MS confirmation.  If the
            Contractor fails to perform GC/MS confirmation as appropriate, the
                                  D-62/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors —  Section  11
                                                Data Analysis and Calculations
                                                    Qualitative Identification
            Agency may require reanalysis of any affected samples  at  no
            additional cost to the Agency.

11.1.2.2    The GC/MS confirmation may be accomplished by one of three general
            means:

            •   Examination of the semivolatile GC/MS library search  results
                (i.e., TIC data), or

            •   A second analysis of the semivolatile extract, or

            •   Analysis of the pesticide/Aroclor extract, following  any
                solvent exchange and concentration steps that may  be
                necessary.

11.1.2.3    The semivolatile GC/MS analysis procedures outlined in Exhibit D
            SVGA are based on the injection into the instrument of
            approximately 20 ng of a target compound in a 2 tiL volume.  The
            semivolatile CRQL values in Exhibit C are based on the sample
            concentration that corresponds to an extract concentration of
            10 ng//xL of target analyte.  However, these are quantitation
            limits, and the detection of analytes and generation of
            reproducible mass spectra will routinely be possible at levels 3-
            10 times lower.  The sample concentration corresponding to
            10 ng//xL in extract will depend on the sample matrix.

11.1.2.3.1     For water samples, 20 ng/2 /xL corresponds to a sample
               concentration of 10 /xg/L.

11.1.2.3.2     For soil/sediment samples prepared according to the
               semivolatile low level soil/sediment method (i.e.,  30  g of
               soil/sediment), the corresponding sample concentration is
               330 /xg/Kg.

11.1.2.3.3     For soil/sediment samples prepared according to the
               semivolatile medium level soil/sediment method (i.e.,  1 g of
               soil/sediment), the corresponding sample concentration is
               10,000 /xg/Kg.

11.1.2.3.4     Therefore,  based on the values given above, any pesticide
               sample in which compound concentration in the sample extract is
               greater than or equal to 10 ng//xL for single component
               pesticides, 50 ng//xL for Aroclors, and 125 ng/iiL for Toxaphene
               should enable the laboratory to confirm the pesticide/Aroclor
               by GC/MS analysis of the semivolatile extract.

11.1.2.4    In order to confirm the identification of the target
            pesticide/Aroclor, the laboratory must also analyze a  reference
            standard for the analyte.  In order to demonstrate the ability of
            the GC/MS system to identify the analyte in question,  the
            concentration of the standard should be 10 ng/iiL for single
            component pesticides, 50 ng//xL for Aroclors, and 125 ng//xL for
            Toxaphene.

                                  D-63/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 11
Data Analysis and Calculations
Qualitative Identification
11.1.2.5    To facilitate the confirmation of the single component pesticide
            analytes from the semivolatile library search data, the laboratory
            may wish to include these analytes in the semivolatile continuing
            calibration standard at a concentration of 10 ng//iL or less.  Do
            not include the Aroclors and toxaphene mixture in the semivolatile
            initial and continuing calibration standard.  If added to this
            GC/MS standard, the response factors, retention times, etc. for
            these analytes would be reported on the GC/MS quantitation report,
            but not on the GC/MS calibration data reporting forms.  As only a
            single concentration of each analyte would be analyzed, no
            linearity (%RSD) or percent difference criteria would be applied
            to the response factors for these additional analytes.

11.1.2.6    The laboratory is advised that library search results from the
            NIST/EPA/NIH (May 1992 release or later) and Wiley (1991 release
            or later) mass spectral library will not likely list the name of
            the pesticide/Aroclor analyte as it appears in this SOW, hence,
            the mass spectral interpretation specialist is advised to compare
            the CAS Registry numbers for the pesticides/Aroclors to those from
            the library search routine.

11.1.2."?    If the analyte cannot be confirmed from the semivolatile library
            search data for the original semivolatile GC/MS analysis, the
            laboratory may analyze another aliquot of the semivolatile sample
            extract after further concentration of the aliquot.  This second
            aliquot must either be analyzed as part of a routine semivolatile
            GC/MS analysis, including instrument performance checks (DFTPP),
            calibration standards containing the pesticides/Aroclors as
            described in Section 11.1.2.5, or it must be analyzed along with
            separate reference standards for the analytes to be confirmed.

11.1.2.8    If the analyte cannot be confirmed by either the procedures in
            Sections 11.1.2.5 or 11.1.2.7, then an aliquot of the extract
            prepared for the GC/EC analysis must be analyzed by GC/MS,
            following any necessary solvent exchange and concentration steps.
            As in Section 11.1.2.%, analysis of a reference standard is
            required if the GC/MS continuing calibration standard does not
            contain the analyte to be confirmed.

11.1.2.9    Regardless of which of the three approaches above is used for
            GC/MS confirmation, the appropriate blank must also be analyzed by
            GC/MS to demonstrate that the presence of the analyte was not the
            result of laboratory contamination.  If the confirmation is based
            on the analysis of the semivolatile extract, then the semivolatile
            method blank extracted with the sample must also be analyzed.  If
            the confirmation is based on the analysis of the extract prepared
            for the GC/EC analysis, then the pesticide method blank extracted
            with the sample must be analyzed.

11.1.2.10   If the identification of the analyte cannot be confirmed by any of
            the GC/MS procedures above and the concentration calculated from
            the GC/EC analysis is greater than or equal to the concentration
            of the reference standard analyzed by GC/MS, then report the

                                   D-64/PEST                           OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
            analyte as undetected, adjust the sample quantitation limit (the
            value associated with the "U" qualifier) to a sample concentration
            equivalent to the concentration of the GC/MS reference standard,
            and qualify the results on Form I with one of the laboratory-
            defined qualifiers ("X," "Y," or "Z").  In this instance, define
            the qualifier explicitly in the SDG Narrative, and describe the
            steps taken to confirm the analyte in the SDG Narrative.

11.1.2.11   For GC/MS confirmation of single component analytes, the required
            deliverables are copies of the library search results (best TIC
            matches) or analyte spectrum and the spectrum of the reference
            standard.  For multicomponent analytes, spectra of three
            characteristic peaks are required for both the sample component
            and the reference standard.

11.1.2.12   The purpose of the GC/MS analysis for the single component
            pesticides is for identification.  The purpose of the GC/MS
            analysis for the multicomponent analytes is to confirm the
            presence of chlorinated biphenyls in Aroclor and the presence of
            chlorinated camphenes in Toxaphene.   The GC/MS analytical results
            for the pesticides/Aroclors shall not be used for quantitation and
            the GC/MS results shall not be reported on Form I and Form X.  The
            exception noted in Section 11.1.2.10 applies only to analytes that
            cannot be confirmed above the reference standard concentration.

11.2  Calculations

11.2.1   Target Compounds

         The concentrations of the single component pesticides and surrogates
         are calculated separately for both GC columns by using the following
         equations.

11.2.1.1    Water
            EQ. 13
                                     .     (Ax) (Ve] (£>f) (GPC]
                    Concentration pg/L =  -*.-*-
                                            (CF) (V0) (Vf)


            Where,

            Ax =  Area of the peak for the compound to be measured.
            CF =  Calibration factor from the initial calibration for the
                  midpoint concentration external standard (area per ng).
            V0 =  Volume of water extracted in milliliters (mL).
            Vj =  Volume of extract injected in microliters (ju.L) .  (If a
                  single injection is made onto two columns,  use one half the
                  volume in the syringe as the volume injected onto each
                  column.)
                                   D-65/PEST                           OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 11
Data Analysis  and Calculations
Calculations
            Vt =   Volume of the concentrated extract in  microliters
                   (If GPC is not performed, then Vt  =  10,000 /iL.   If GPC is
                   performed, then Vt = 5,000 /iL. )
            Df =   Dilution factor.  The dilution factor for analysis of water
                   samples by this method is defined  as follows:

                    iA. most cone, extract used to make dilution * ^L clean  solvent
                         ML most cone, extract used to make dilution

                   If  no dilution is performed,  Df  =  1.0.

            GPC =  GPC factor.  (If no GPC is performed, GPC =1.   If GPC is
                   performed, then GPC = 2.0)

11.2.1.2    Soil/Sediment
            EQ.  14

                            ,„  ,„      • i,  i,   • \    (Ax) (Vt) (Df} (GPC)
           Concentration pg/Kg (Dry weight basis] =
                                                      (CF) (VJ (fig (D)
            Where,

            Ax and CF are  as  given for water, above.
            Vt =   5,000 /iL.
            Vj =   Volume of  extract injected in microliters (/*L) .    (If a
                   single injection is made onto two  columns, use one half the
                   volume in  the syringe as the volume  injected onto each
                   column.)
             D =   100  - %  moisture
                         100
            Ws =   Weight of  sample extracted in grams  (g)
            Df =   Dilution factor.  The dilution  factor for analysis of
                   soil/sediment samples by this method is  defined as follows:

                    nl. most cone, extract used to make dilution + y.L clean solvent
                         j»L most cone, extract used to make dilution

                   If no dilution is performed, Df =  1.0.

            GPC =  GPC  factor = 2.

11.2.1.2.1     The GPC  factor is used to account  for the amount of  extract
               that  is  not recovered from the mandatory use of GPC  cleanup.
               Concentrating the extract collected after GPC to 5.0 mL rather
               than  10.0 mL  for water samples not subjected to GPC maintains
               the sensitivity of the soil/sediment  method comparable to that
               of  the  water  method, but correction of  the  numerical results  is
               still required.
                                    D-66/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 11
                                                Data Analysis and Calculations
                                                                  Calculations
11.2.1.2.2     Note that the calibration factors used for the quantitation of
               the single component pesticides are the calibration factors
               from the midpoint concentration standard in the most recent
               initial calibration.

11.2.1.2.3     Because of the likelihood that compounds co-eluting with the
               target compounds will cause positive interferences and increase
               the concentration determined by the method, the lower of the
               two concentrations calculated for each single component
               pesticide is reported on Form I.  In addition, the
               concentrations calculated for both the GC columns are reported
               on Form X, along with a percent difference comparing the two
               concentrations.  The percent difference is calculated according
               to Equation 15.
               EQ. 15
ConcH - ConcL
    ConcL
                                              x 100
               Where,

               ConcH =  The higher of the two concentrations
                        for the target compound in question
               ConcL =  The lower of the two concentrations
                        for the target compound in question

11.2.1.2.4     Note that using this equation will result in percent difference
               values that are always positive.  The value will also be
               greater than a value calculated using the higher concentration
               in the denominator; however, given the likelihood of a positive
               interference raising the concentration determined on one GC
               column, this is a conservative approach to comparing the two
               concentrat ions.

11.2.1.2.5     The quantitative determination of Toxaphene or Aroclors is
               somewhat different from that of single component pesticides.
               Quantitation of peaks within the detector linear range CRQL to
               > 16 times CRQL is based on a single calibration point assuming
               linear detector response.   Alternatively, a linear calibration
               range may be established during a run sequence by a three-point
               calibration curve for any multicomponent analyte.  If the
               concentration is calculated to be 106 times  the CRQL,  the
               Contractor shall contact SMO immediately.

11.2.1.2.6     The quantitation of toxaphene or Aroclors must be accomplished
               by comparing the heights or the areas of each of the three to
               five major peaks of the multicomponent analyte in the sample
               with the calibration factor for the same peaks established
               during the initial calibration sequence.  The concentration of
               multicomponent analytes is calculated by using Equations 13 and
               14, where Ax is  the area for each of the major peaks of the


                                   D-67/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors  —  Section  11
Data Analysis and Calculations
Calculations
               multicomponent  analyte.  The concentration  of  each  peak is
               determined and  then a mean concentration  for the  three to five
               major peaks is  determined on each column.

11.2.1.2.7     The reporting requirements for Toxaphene  and the  Aroclors are
               similar to those for the single component analytes,  except that
               the lower mean  concentration (from three  to five  peaks)  is
               reported on Form I, and the two mean concentrations reported on
               Form X.  The two mean concentrations are  compared by
               calculating the percent difference using  Equation 15.

11.2.2   CRQL Calculation

         If the adjusted CRQL  is less than the CRQL listed in Exhibit C
         (Pesticides), report  the CRQL in Exhibit C (Pesticides).

11.2.2.1    Water Samples

      EQ. 16

                     Adjusted _ Contract   (Vx> (Vtl (fy
                       CRQL   '  CRQL   X   (VQ) ( vg (
            Where,

            Vt,  Df,  V0,  and  V,- are as given  in equation  13.
            Vx =  Contract sample volume (1000 mL).
            Vy =  Contract injection volume (1 jiL or 2 /iL) .
            Vc =  Contract concentrated extract volume  (10,000 /*L  if GPC was
                  not performed and 5,000 jtL if GPC was performed.
11.2.2.2    Soil/Sediment Samples

      EQ. 17
                     Adjusted = Contract    (Wx) (Vt) (Vy) (Df)
                       CRQL      CRQL      (W) (y  (V.) (D)
            Where,
            Vt,  Df,  Ws,  V; and D are as given in equation  14.
            Wx =  Contract sample weight (30 g).
            Vy =  Contract injection volume (1 n~L or 2 j*L) .
            Vc =  Contract concentrated extract volume (GPC is required:
                  5000 yxL) .

11.2.3   Surrogate Recoveries

11.2.3.1    The concentrations of the surrogates are calculated  separately for
            each GC column in a similar manner as the other analytes,  using
            Equations 13 and 14.  Use the calibration factors from the
            midpoint concentration of Individual Standard  Mixture  A from the
            initial calibration.  The recoveries of the surrogates are

                                   D-68/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 11
                                                Data Analysis and Calculations
                             Technical Acceptance Criteria for Sample Analysis
             calculated  for each GC column according to Equation  12, repeated
             below.
            EQ.  12
            Where,
                                Percent Recovery - —^ x 100
            Qd = Quantity determined by analysis
            Qa = Quantity added
11.2.3.2    The advisory  limits for the recovery of the surrogates are 30 -
            150 percent for both surrogate compounds.

11.2.3.3    As these limits are only advisory, no further action is required
            by the laboratory; however, frequent failures to meet the limits
            for surrogate recovery warrant investigation by the laboratory,
            and may result in questions from the Agency.  Surrogate recovery
            data from both GC columns are reported  (see Exhibit B).

11.3  Technical Acceptance Criteria for Sample Analysis

      The requirements below apply independently ^o each GC column and to all
      instruments used for these analyses.  Quantstation must be performed on
      each GC column.

11.3.1   Samples must be analyzed under the GC/EC operating  conditions in
         Section 9.  The instrument must have met all initial calibration,
         calibration verification,  and blank technical acceptance criteria.
         Samples must be cleaned-up, when required, on a GPC meeting the
         technical acceptance criteria for GPC calibration and GPC calibration
         checks.  Samples must be cleaned-up using florisil meeting the
         technical acceptance criteria for florisil.  Sample data must be
         bracketed at 12-hour intervals (or less) by acceptable analyses of
         instrument blanks, PEMs and Individual Standard Mixtures A and B, as
         described in Section 10.2.2.1.

11.3.2   The samples must be extracted and analyzed within the contract
         required holding times.

11.3.3   The samples must have an associated method blank meeting the
         technical acceptance criteria for method blanks.  When sulfur cleanup
         blanks are required, the samples must have associated with it a
         sulfur cleanup blank meeting the technical acceptance criteria for
         sulfur cleanup blanks.

11.3.4   The retention time for each of the surrogates must be within the
         retention time window as calculated in Section 9 for both GC columns.
                                   D-69/PEST                            OLM03.0

-------
Exhibit D Pesticide/Aroclors — Section 11
Data Analysis and Calculations
Technical Acceptance Criteria for Sample Analysis
 11.3.5   No target analyte concentrations may exceed the upper limit of the
         initial calibration, or else the extract must be diluted and
         reanalyzed.

 11.3.6   A standard for any identified multicomponent analyte must be analyzed
         during a valid analytical sequence on the same instrument and column,
         within 72 hours of its detection in a sample.

 11.3.7   The identification of single component pesticides by gas
         chromatographic methods is based primarily on retention time data.
         The retention time of the apex of a peak can be verified only from an
         on-scale chromatogram.  The identification of multicomponent analytes
         is based primarily on recognition of patterns of retention times
         displayed on a chromatogram.  Therefore, the following requirements
         apply to all data presented for single component and multicomponent
         analytes.

 11.3.7.1    When no analytes are identified in a sample,  the chromatograms
            from the analyses of the sample extract must use the same scaling
            factor as was used for the low point standard of the initial
            calibration associated with those analyses.

 11.3.7.2    Chromatograms must display single component pesticides detected in
            the sample at less than full scale.

 11.3.7.3    Chromatograms must display the largest peak of any multicomponent
            analyte detected in the sample at less than full scale.

 11.3.7.4    If an extract must be diluted, chromatograms must display single
            component pesticides between 10 and 100 percent of full scale.

 11.3.7.5    If an extract must be diluted, chromatograms must display the
            peaks chosen for quantitation of multicomponent analytes between
            25 and  100 percent of full scale.

11.3.7.6    For any sample or blank,  the baseline of the chromatogram must
            return  to below 50 percent of full scale before the elution time
            of alpha-BHC,  and return to below 25 percent of full scale after
            the elution time of alpha-BHC and before the elution time of
            decachlorobiphenyl.

11.3.7.7    If a chromatogram is replotted electronically to meet these
            requirements,  the scaling factor used must be displayed on the
            chromatogram.

11.3.7.8    If the  chromatogram of any sample needs to be replotted
            electronically to meet these requirements,  both the initial
            chromatogram and the replotted chromatogram must be submitted in
            the data package.
                                  D-70/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 11
                                                Data Analysis and Calculations
                                         Corrective Action for Sample Analysis
11.4  Corrective Action for Sample Analysis

11.4.1   Sample analysis technical acceptance criteria MUST be met before data
         are reported.  Samples contaminated from laboratory sources or
         associated with a contaminated method blank or sulfur cleanup blank
         will require re-extraction and reanalysis at no additional cost to
         the Agency.  Any samples analyzed that do not meet the technical
         acceptance criteria will require re-extraction and/or reanalysis at
         no additional cost to the Agency.

11.4.2   If the sample analysis technical acceptance criteria are not met,
         check calculations, surrogate solutions, and instrument performance.
         It may be necessary to recalibrate the instrument or take other
         corrective action procedures to meet the technical acceptance
         criteria, in which case, the affected samples must be reanalyzed at
         no additional cost to the Agency after the corrective action.

11.4.3   The extract from samples which were cleaned-up by GPC using an
         automated injection system and have surrogate recoveries outside the
         lower advisory surrogate acceptance limits must be checked to assure
         that the proper amount was injected on the GPC column.   If
         insufficient volume was injected, the sample must be reprepared and
         reanalyzed at no additional cost to the Agency.

11.4.4   If sample chromatograms have a high baseline or interfering peaks,
         inspect the system to determine the cause of the problem (e.g.
         carryo er, column bleed, dirty BCD, contaminated gasses, leaking
         septum, etc.).   After correcting the problem, analyze an instrument
         blank to demonstrate that the system is functioning properly.
         Reanalyze the sample extracts.  If the problem with the samples still
         exists, then those samples must be re-extracted and reanalyzed.
         Samples which cannot be made to meet the given specifications after
         one re-extraction and three-step cleanup (GPC,  Florisil, and sulfur
         cleanups) are reported in the SDG Narrative and do not  require
         further analysis.
                                  D-71/PEST                            OLM03.1

-------
Exhibit D Pesticides/Aroclors — Section 12
Quality Control
Blank Analyses
 12.0  QUALITY CONTROL

 12.1  Blank Analyses

 12.1.1   Introduct ion

         There are two types of blanks always required by this method:  the
         method blank and the instrument blank.  A separate sulfur cleanup
         blank may be required if all samples associated with a given method
         blank are not subjected to sulfur cleanup.  Samples that are
         associated with a sulfur cleanup blank are also associated with the
         method blank with which they were extracted.  Both the method and
         sulfur cleanup blanks must meet the respective acceptance criteria
         for the sample analysis acceptance criteria to be met.

 12.1.2   Method Blanks

 12.1.2.1    Summary of Method Blanks

            A method blank is a volume of a clean reference matrix (reagent
            water for water samples, or purified sodium sulfate for
            soil/sediment samples) that is carried through the entire
            analytical procedure.  The volume or weight of the reference
            matrix must be approximately equal to the volume or weight of
            samples  associated with the blank.  The purpose of a method blank
            is to determine the levels of contamination associated with the
            processing and analysis of samples.

12.1.2.2    Frequency of Method Blanks

            A method blank must be extracted once for the following,  whichever
            is most frequent,  and analyzed on each GC/EC system used to
            analyze samples:

            •   Each SDG (not to exceed 20 field samples),  or

            •   Each matrix within an SDG, or

            •   Each extraction procedure within an SDG, or

            •   Whenever samples are extracted.

12.1.2.3    Procedure for Method Blank Preparation

12.1.2.3.1      For pesticide/Aroclor analyses,  a method blank for water
               samples consists of a 1 L volume of reagent water spiked with
               I mL of the surrogate spiking solution (Section 7.2.4.1).  For
               soil/sediment  samples,  the method blank consists of 30 g of
               sodium sulfate spiked with 2 mL of the surrogate spiking
               solution.

12.1.2.3.2      Extract,  concentrate and analyze method blanks according to
               Section 10.

                                  D-72/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 12
                                                               Quality Control
                                                                Blank Analyses
12.1.2.3.3     Calculate method blank results according to Section 11.

12.1.2.4    Technical Acceptance Criteria for Method Blanks

12.1.2.4.1     The requirements below apply independently to each GC column
               and to all instruments used for these analyses.  Quantitation
               must be performed on both GC columns.

12.1.2.4.2     All method blanks must be prepared and analyzed at the
               frequency described in Section 12.1.2.2 using the procedure
               above and in Section 10 on a GC/EC system meeting the initial
               calibration and calibration verification technical acceptance
               criteria.  Method blanks must undergo cleanup, when required,
               on a GPC meeting the technical acceptance criteria for GPC
               calibration and GPC calibration checks.  Method blanks must be
               cleaned-up using Florisil meeting the technical acceptance
               criteria for Florisil.  Method blanks must be bracketed at 12-
               hour intervals (or less) by acceptable analyses of instrument
               blanks, PEMS, and individual standard mixtures A and B as
               described in Section 10.2.2.1.

12.1.2.4.3     The concentration of the target compounds (Exhibit C
               (Pesticides)) in the method blank must be less than the CRQL
               for each target compound.

12.1.2.4.4     The method blank must meet all sample technical acceptance
               criteria in Sections 11.3.4 to 11.3.7

12.1.2.4.5     Surrogate recoveries must fall within the acceptance windows of
               30-150%.  In the case of the method blank(s), these limits are
               not advisory.

12.1.2.5    Corrective Action for Method Blanks

12.1.2.5.1     If a method blank does not meet the technical acceptance
               criteria, the Contractor must consider the system to be out of
               control.

12.1.2.5.2     If contamination is a problem, then the source of the
               contamination must be investigated and appropriate corrective
               measures must be taken and documented before further sample
               analysis proceeds.  It is the Contractor's responsibility to
               ensure that method interferences caused by contaminants in
               solvents, reagents, glassware, and sample storage and sample
               processing hardware that lead to discrete artifacts and/or
               elevated baselines be investigated and appropriate corrective
               actions be taken and documented before further sample analysis.
               All samples associated with a contaminated method blank must be
               re-extracted/reanalyzed at no additional cost to the Agency.

12.1.2.5.3     If the surrogate recoveries in the method blank do not meet the
               acceptance criteria listed in 12.1.2.4.5, first reanalzye the
               method blank.  If surrogate recoveries do not meet the

                                   D-73/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 12
Quality Control
Blank Analyses
               acceptance criteria after reanalysis,  the method blank and all
               samples associated with that method blank must  be re-extracted
               and reanalyzed at no additional cost to the Agency.

12.1.2.5.4     If the method blank failed to meet  the criteria listed in
               Sections 12.1.2.4.2 and 12.1.2.4.4, then there  is an instrument
               problem.  Correct the instrument problem and reanalyze the
               method blank.

12.1.3   Sulfur Cleanup Blanks

12.1.3.1    Summary of Sulfur Cleanup Blanks

            The sulfur cleanup blank is a modified form of the method blank.
            The sulfur cleanup blank is hexane spiked with the surrogates and
            carried through the sulfur cleanup and analysis procedures.   The
            purpose of the sulfur cleanup blank is to determine the levels of
            contamination associated with the separate sulfur  cleanup steps.

12.1.3.2    Frequency of Sulfur Cleanup Blanks

            The sulfur cleanup blank is prepared separately when only part of
            a set of samples extracted together requires sulfur removal.   A
            method blank is associated with the entire set of  samples.   The
            sulfur cleanup blank is associated with the part of the set  which
            required sulfur cleanup.  If all the samples associated with a
            given method blank are subjected to sulfur cleanup, then the
            method blank must be subjected to sulfur cleanup,  and no separate
            sulfur cleanup blank is required.

12.1.3.3    Procedure for Sulfur Cleanup Blank

12.1.3.3.1     The concentrated volume of the blank must be the same as  the
               final volume of the samples associated with the blank.  The
               sulfur blank must also contain the  surrogates at the same
               concentrations as the sample extracts (assuming 100.0 percent
               recovery).  Therefore, add 0.1 mL of the surrogate solution to
               0.9 mL of hexane in a clean vial, or for a sulfur blank with a
               final volume of 2 mL, add 0.2 mL of the surrogate solution to
               1.8 mL of hexane in a clean vial.

12.1.3.3.2     Proceed with the sulfur removal (Section 10.1.8.3.3.1 or
               10.1.8.3.3.2) using the same technique (mercury or copper) as
               the samples associated with the blank.

12.1.3.3.3     Analyze the sulfur cleanup blank according to Section 10.2.
               Assuming that the material in the sulfur cleanup blank resulted
               from the extraction of a 1 L water  sample, calculate the
               concentration of each analyte using the equation in Section
               11.2.1.1.  Compare the results to the CRQL values for water
               samples in Exhibit C (Pesticides).
                                  D-74/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 12
                                                               Quality Control
                                                                Blank Analyses
12.1.3.4    Technical Acceptance Criteria For Sulfur Cleanup Blanks

12.1.3.4.1     The requirements below apply independently to each GC column
               and to all  instruments used for these analyses.  Quantitation
               must be performed on both GC columns.

12.1.3.4.2     All sulfur  cleanup blanks must be prepared and analyzed at the
               frequency described in Section 12.1.3.2 using the procedure
               referenced  in Section 12.1.3.3 on a GC/EC system meeting the
               initial calibration and calibration verification technical
               acceptance  criteria.

12.1.3.4.3     Sulfur cleanup blanks must be bracketed at 12-hour intervals
               (or less) by acceptable analyses of instrument blanks, PEMs,
               and Individual Standard Mixtures A and B, as described in
               Section 10.2.2.1.

12.1.3.4.4     The concentration of the target compounds (Exhibit C
               (Pesticides)) in the sulfur cleanup blank must be less than the
               CRQL for each target compound.

12.1.3.^.5     The sulfur  cleanup blank must meet all sample technical
               acceptance  criteria in Sections 11.3.4 to 11.3.7.

12.1.3.4.6     Surrogate recoveries must fall within the acceptance windows of
               30-150%.  In the case of the sulfur cleanup blank, these limits
               are not advisory.

12.1.3.5    Corrective Action for Sulfur Cleanup Blanks

12.1.3.5.1     If a sulfur cleanup blank does not meet the technical
               acceptance  criteria, the Contractor must consider the system to
               be out of control.

12.2.3.5.2     If contamination is a problem, then the source of the
               contamination must oe investigated and appropriate corrective
               measures must be taken and documented before further sample
               analysis proceeds.   It is the Contractor's responsibility to
               ensure that method interferences caused by contaminants in
               solvents,  reagents,  glassware, and sample storage and sample
               processing  hardware that lead to discrete artifacts and/or
               elevated baselines be investigated and appropriate corrective
               actions be taken and documented before further sample analysis.
               All samples associated with a contaminated sulfur cleanup blank
               must be re-extracted/reanalyzed at no additional cost to the
               Agency.

12.1.3.5.3     If the surrogate recoveries in the sulfur cleanup blank do not
               meet the acceptance criteris listed in section 12.1.3.4.5,
               first reananlyze the sulfur cleanup blank.  If surrogate
               recoveries do not meet the acceptance criteria after
               reanalysis,  the sulfur cleanup blank and all samples associated
                                   D-75/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 12
Quality Control
Blank Analyses
               with that sulfur cleanup blank must be re-extracted and
               reanalyzed at no additional cost to the Agency.

12.1.3.5.4     If the sulfur cleanup blank failed to meet the criteria used in
               12.1.3.4.2 and 12.1.3.4.5, then there is an instrument problem.
               Correct the instrument problem and reanalyze the sulfur cleanup
               blank.

12.1.4   Instrument Blanks

12.1.4.1    Summary of Instrument Blanks

            An instrument blank is a volume of clean solvent spiked with the
            surrogates and analyzed on each GC column and instrument used for
            sample analysis.  The purpose of the instrument blank is to
            determine the levels of contamination associated with the
            instrumental analysis itself, particularly with regard to the
            carry over of analytes from standards or highly contaminated
            samples into other analyses.

12.1.4.2    Frequency of Instrument Blanks

            The first analysis in a 12-hour analysis sequence must be an
            instrument blank.  All acceptable sample analyses are to be
            bracketed by acceptable instrument blanks, as described in Section
            10.2.2.1  If more than 12 hours have elapse 1 since the injection
            of t le instrument blank that bracketed a previous 12-hour period,
            an instrument blank must be analyzed to initiate a new 12-hour
            sequence.

12.1.4.3    Procedure for Instrument Blanks

12.1.4.3.1     Prepare the instrument blank by spiking the surrogates into
               hexane or iso-octane for a concentration of 20 ng/mL of
               tetrachloro-m-xylene and decachlorobiphenyl.

12.1.4.3.2     Analyze the instrument blank according to Section 10.2 at the
               frequency listed in Section 12.1.4.2

12.1.4.3.3     For comparing the results of the instrument blank analysis to
               the CRQLs, assume that the material in the instrument resulted
               from the extraction of a 1 L water sample and calculate the
               concentration of each analyte using the equation in Section
               11.2.1.1.  Compare the results to one-half the CRQL values for
               water samples in Exhibit C (Pesticides).

12.1.4.4    Technical Acceptance Criteria for Instrument Blanks

12.1.4.4.1     All instrument blanks must be prepared and analyzed at the
               frequency described in Section 12.1.4.2 using the procedure in
               Section 12.1.4.3 on a GC/EC system meeting the initial
               calibration and calibration verification technical acceptance
               criteria.

                                   D-76/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 12
                                                               Quality Control
                                                                        MS/MSD
12.1.4.4.2     The concentration of each of the target analytes (Exhibit C
               (Pesticides)) in the instrument blank must be less than 0.5
               times the CRQL for that analyte.

12.1.4.4.3     The instrument blank must meet all sample technical acceptance
               criteria in Section 11.3.4 to 11.3.7.

12.1.4.5    Corrective Action for Instrument Blanks

12.1.4.5.1     If analytes are detected at greater than half the CRQL, or the
               surrogate RTs are outside the RT windows, all data collection
               must be stopped, and corrective action must be taken.  Data for
               samples which were run between the last acceptable instrument
               blank and the unacceptable blank are considered suspect.  An
               acceptable instrument blank must be run before additional data
               are collected.  After an acceptable instrument blank is run,
               all samples which were considered suspect as defined by the
               criteria described above must be reinjected during a valid run
               sequence at no additional cost to the Agency and must be
               reported.
                                  D-77/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 12
Quality Control
MS/MSD
12.2  Matrix Spike/Matrix Spike Duplicate (MS/MSD)

12.2.1   Summary of MS/MSD

         In order to evaluate the effects of the sample matrix on the methods
         used for pesticide/Aroclor analyses, the Agency has prescribed a
         mixture of pesticide/Aroclor target compounds to be spiked into two
         aliquots of a sample,  and analyzed in accordance with the appropriate
         method.

12.2.2   Frequency of MS/MSD Analysis

12.2.2.1    A matrix spike and matrix spike duplicate must be extracted and
            analyzed at least every 20 samples of each matrix.  NOTE:  There
            is no differentiation between "low" and "medium" level
            soil/sediment samples in this method.   Therefore only one
            soil/sediment MS/MSD is to be submitted per Sample Delivery Group
            (SDG).

12.2.2.2    As part of the Agency's QA/QC program,  water rinsate samples
            and/or field blanks may be delivered to a laboratory for analysis.
            Do not perform MS/MSD analysis on a water rinsate sample or field
            blank.

12.2.2.3    If the EPA Region designates a sample to be used as an MS/MSD,
            then that sample must be used.  If there is insufficient sample
            volume remaining to perform an MS/MSD,  then the Contractor shall
            choose another sample to perform an MS/MSD analysis.  At the time
            the selection is made,  the Contractor shall notify the Region
            (through SMO) that insufficient sample  was received and identify
            the EPA sample selected for the MS/MSD  analysis.  The rationale
            for the choice of another sample other  than the one designated by
            the Agency shall be documented in the SDG Narrative.

12.2.2.4    If there is insufficient sample volume  remaining in any of the
            samples in an SDG to perform an MS/MSD, the Contractor shall
            immediately contact SMO to inform them  of the problem.  SMO will
            contact the Region for instructions. The Region will either
            approve that no MS/MSD be performed, or require that a reduced
            sample aliquot be used for the MS/MSD analysis.  SMO will notify
            the Contractor of the Region's decision.  The Contractor shall
            document the decision in the SDG Narrative.

12.2.2.5    The Contractor will not be paid for MS/MSD analysis performed at a
            greater frequency then required by the  contract.  If it appears
            that the Region has requested MS/MSD analysis at a greater
            frequency then required by the contract, the Contractor shall
            contact SMO.  SMO will contact the Region to determine which
            samples should have an MS/MSD performed on them.  SMO will notify
            the Contractor of the Region's decision.  The Contractor shall
            document the decision in the SDG Narrative.
                                  D-78/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 12
                                                               Quality Control
                                                                        MS/MSD
 12.2.2.6    When  a Contractor receives only performance evaluation  (PE)
            samples,  no MS/MSD  shall be performed within that SDG.

 12.2.2.7    When  a Contractor receives a PE sample as part of a larger SDG, a
            sample other than the PE sample must be chosen for the  MS/MSD when
            the Region did not  designate samples to be used for this purpose.
            If the PE sample is received as an ampulated standard extract, the
            ampulated PE sample is not considered to be another matrix type.

 12.2.3   Procedure for Preparing MS/MSD

 12.2.3.1    Water Samples

            For water samples,  measure out two additional 1 L aliquots of the
            sample chosen for spiking.  Adjust the pH of the samples (if
            required) and fortify each with 1 mL of matrix spiking  solution.
            Using a syringe or  volumetric pipet, add 1 mL of surrogate spiking
            solution to each sample.  Extract, concentrate, cleanup, and
            analyze matrix spikes and matrix spike duplicate according to
            Section 10.0.

 12.2.3.2    Soil/Sediment Samples

            For soil/sediment samples weigh out two additional 30 g (record
            weight to the nearest 0.1 g)  aliguots of the sample chosen for
            spiking.   Add 1 mL  of matrix ppiking solution and 2 mL of
            surrogate solution.   Extract,  concentrate, cleanup, and analyze
            matrix spikes and matrix spike duplicates according to Section
            10.0.

12.2.3.3    Note:   Before any MS/MSD analysis, analyze the original sample,
            then analyze the MS/MSD at the same concentration as the most
            concentrated extract for which the original sample results will be
            reported.  For example, if the original sample is to be reported
            at a 1:1  dilution and a 1:10 dilution,  then analyze and report the
            MS/MSD at a 1:1 dilution only.   However,  if the original sample is
            to be reported at a 1:10 dilution and a 1:100 dilution, then the
            MS/MSD must be analyzed and reported at a 1:10 dilution only.  Do
            not further dilute the MS/MSD  samples to get either spiked or non-
            spiked analytes within calibration range.

12.2.4   Calculations for MS/MSD

12.2.4.1    The percent recoveries and the relative percent difference between
            the recoveries of each of the  compounds in the matrix spike
            samples will be calculated and reported by using the following
            equations:
                                  D-79/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 12
Quality Control
MS/MSD
            EQ. 18
                     Matrix Spike Recovery = SSR ~ SR x 100
                                               SA
            Where,
            SSR =  Spike sample result
            SR  =  Sample result
            SA  =  Spike added
EQ. 19
                          - MSDR\
                           RPD =       -        x 100
                                   (MSR + MSDR]
            Where,

            RPD =  Relative percent difference
            MSR =  Matrix spike recovery
            MSDR=  Matrix spike duplicate recovery

12.2.4.2    The vertical bars in the formula above indicate the absolute value
            of the difference,  hence RPD is always expressed as a positive
            value.

12.2.5   Technical Acceptance Criteria for MS/MSD

12.2.5.1    The requirements below apply independently to each GC column and
            to all instruments used for these analyses.   Quantitation must be
            performed on both GC columns.

12.2.5.2    All MS/MSD must be prepared and analyzed at the frequency
            described in Section 12.2.2 using the procedure above and in
            Section 10 on a GC/EC system meeting the initial calibration,
            calibration verification,  and blank technical acceptance criteria.
            MS/MSD must be cleaned-up,  when required,  on a GPC meeting the
            technical acceptance criteria for GPC calibration and GPC
            calibration checks.   MS/MSD must be cleaned-up using florisil
            meeting the technical acceptance criteria for florisil.  MS/MSD
            must be bracketed at 12-hour intervals (or less) by acceptable
            analyses of instrument blanks, PEMS, and individual standard
            mixtures A and B as described in Section 10.2.2.1.

12.2.5.3    The samples must be extracted and analyzed within the contract
            required holding times.

12.2.5.4    The retention time for each of the surrogates must be within the
            retention time window as calculated in Section 9 for both GC
            columns.
                                  D-80/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 12
                                                               Quality Control
                                                                        MS/MSD
12.2.5.5    The limits for matrix spike compound recovery and RPD are given in
            Table 3.  As these limits are only advisory, no further action by
            the laboratory is required.  However, frequent failures to meet
            the limits for recovery or RPD warrant investigation by the
            laboratory, and may result in questions from the Agency.

12.2.6   Corrective Action for MS/MSD

         Any MS/MSD which fails to meet the technical acceptance criteria for
         MS/MSD must be reanalyzed at no additional cost on the Agency.
                                  D-81/PEST                            OLM03.0

-------
Exhibit D Pesticides/Aroclors — Sections 13 - 16
Method Performance/Pollution Prevention/Waste Management/References
13.0  METHOD PERFORMANCE

      Not Applicable

14.0  POLLUTION PREVENTION

14.1  Pollution prevention encompasses any technique that reduces or
      eliminates the quantity or toxicity of waste at the point of generation.
      Numerous opportunities for pollution prevention exist in laboratory
      operation.  The EPA has established a preferred hierarchy of
      environmental management techniques that places pollution prevention as
      the management option of first choice.  Whenever feasible, laboratory
      personnel should use pollution prevention techniques to address their
      waste generation.  When wastes cannot be feasibly reduced at the source,
      the Agency recommends recycling as the next best option.

14.2  For information about pollution prevention that may be applicable to
      laboratories and research institutions, consult Less is Better:
      Laboratory Chemical Management for Waste Reduction, available from the
      American Chemical Society's Department of Government Relations and
      Science Policy, 1155 16th Street N.W., Washington,  D.C. 20036, (202)872-
      4477.

15.0  WASTE MANAGEMENT

      The Environmental Protection Agency requires that laboratory waste
      management practices be conducted consistent with all applicable rules
      and regulations.  The Agency urges laboratories to protect the air,
      water,  and land by minimizing and controlling all releases from hoods
      and bench operations, complying with the letter and spirit of any sewer
      discharge permits and regulations, and by complying  with all solid and
      hazardous waste regulations, particularly the hazardous waste
      identification rules and land disposal restrictions.  For further
      information on waste management consult The Waste Management Manual for
      Laboratory Personnel, available from the American Chemical Society at
      the address listed in Section 14.2.

16.0  REFERENCES

      Not Applicable
                                   D-82/PEST                            OLM03.0

-------
                                   Exhibit D Pesticides/Aroclors — Section 17
                                                    Tables/Diagrams/Flowcharts
17.0  TABLES/DIAGRAMS/FLOWCHARTS
                                   Table  1

                    Retention Time Windows for Single and
                    Multicomponent Analytes and Surrogates
            Compound
Retention Time Window (minutes)
alpha-BHC
beta-BHC
gamma-BHC (Lindane)
delta-BHC
Heptachlor
Aldrin
alpha-Chlordane
gamma-Chlordane
Heptachlor epoxide
Dieldrin
Endrin
Endrin aldehyde
Endrin ketone
4, 4 '-ODD
4, 4 '-DDE
4, 4 '-DDT
Endosulfan I
Endosulfan II
Endosulfan sulfate
Met hoxy c h 1 or
Aroclors
Toxaphene
Tetrachloro-m-xylene
Decachlorobiphenyl
±0.05
± 0.05
± 0.05
± 0.05
± 0.05
± 0.05
± 0.07
± 0.07
± 0.07
± 0.07
± 0.07
± 0.07
± 0.07
± 0.07
± 0.07
+ 0.07
±0.07
±0.07
± 0.07
± 0.07
± 0.07
±0.07
± 0.05
±0.10
                                  D-83/PEST
                                     OLM03.0

-------
Exhibit D Pesticides/Aroclors — Section 17
Tables/Diagrams/Flowcharts
                                   Table 2
                    Number of Potential Quantitation Peaks
                                          No.  of Potential
                  Multicomponent Analyte  Quantitation Peaks
                  Aroclor 1016/1260
                  Aroclor 1221
                  Aroclor 1232
                  Aroclor 1242
                  Aroclor 1248
                  Aroclor 1254
                  Toxaphene
                 5/5
                  3
                  4
                  5
                  5
                  5
                  4
                                   Table 3
                         Matrix Spike Recovery and
                      Relative  Percent  Difference  Limits
           Compound
%Recovery  RPD    %Recovery  RPD
Water      Water  Soil       Soil
gamma-BHC (Lindane)
Heptachlor
Aldrin
Dieldrin
Endrin
4, 4 '-DDT
56-123
40-131
40-120
52-126
56-121
38-127
15
20
22
18
21
27
46-127
35-130
34-132
31-134
42-139
23-134
50
31
43
38
45
50
                                  D-84/PEST
                                       OLM03.0

-------
                          EXHIBIT E
QUALITY ASSURANCE/QUALITY CONTROL PROCEDURES AND REQUIREMENTS
                             E-l                               OLM03.0

-------
   Exhibit E - Quality Assurance/Quality Control Procedures and Requirements

                               Table of  Contents


Section                                                                   page

1.0   OVERVIEW	    4

2.0   INTRODUCTION  .....  	    5

3.0   QUALITY ASSURANCE PLAN   	    7
      3.1   Introduction  	    7
      3.2   Required Elements of a Quality Assurance Plan  	    7
      3.3   Updating and Submitting the Quality Assurance Plan  	    9
      3.4   Corrective Actions  	   10

4.0   STANDARD OPERATING PROCEDURES 	   11
      4.1   Introduction	   11
      4.2   Format	   12
      4.3   Requirements	   12
      4.4   Updating and Submitting SOPs	   15
      4.5   Corrective Actions  	   16

5.0   ANALYTICAL STANDARDS REQUIREMENTS 	   18
      5.1   Overview	   18
      5.2   Preparation of Chemical Standards from the Neat High Purity
            Bulk Material	   18
      5.3   Purchase of Chemical Standards Already in Solution  	   19
      5.4   Requesting Standards From the EPA Standards Repository  ...   22
      5.5   Documentation of the Verification and Preparation of
            Chemical Standards  	   22
      5.6   Corrective Actions  	   23

6.0   CONTRACT COMPLIANCE SCREENING 	   24

7.0   REGIONAL DATA REVIEW	   25

8.0   LABORATORY EVALUATION SAMPLES 	   26

9.0   GC/MS TAPE AUDITS	   28
      9.1   Overview	   28
      9.2   Submission of the GC/MS Tape	   29
      9.3   Responding to the GC/MS Tape Audit Report	   30
      9.4   Corrective Actions  	   31

10.0  DATA PACKAGE AUDITS	   32
      10.1  Overview	   32
      10.2  Responding to the Data Package Audit Report	   32
      10.3  Corrective Actions  	   32

11.0  ON-SITE LABORATORY EVALUATIONS  	   33
      11.1  Overview	   33
      11.2  Quality Assurance On-Site Evaluation  	   33
      11.3  Evidentiary Audit  	   33

                                     E-2                               OLM03.0

-------
      11.4'  Discussion of the On-Site Team's Findings 	   34
      11.5  Corrective Action Reports for Follow-Through to Quality
            Assurance and Evidentiary Audit Reports 	   34
      11.6  Corrective Actions  	   35

12.0  QUALITY ASSURANCE AND DATA TREND ANALYSIS 	   36

13.0  DATA MANAGEMENT	   37
                                     E-3                               OLM03.0

-------
Exhibit E — Section 1
Overview
1.0   OVERVIEW

1.1   Quality assurance and quality control are integral parts of the
      Environmental Protection Agency's (EPA)  Contract Laboratory Program
      (CLP).  The quality assurance (QA)  process consists of management review
      and oversight at the planning,  implementation,  and completion stages of
      the environmental data collection activity,  and ensures that data
      provided are of the quality required.  The quality control (QC) process
      includes those activities required during data  collection to produce the
      data quality desired and to document the quality of the collected data.

1.2   During the planning of an environmental  data collection program, QA
      activities focus on defining data quality criteria and designing a QC
      system to measure the quality of data being generated.  During the
      implementation of the data collection effort, QA activities ensure that
      the QC system is functioning effectively,  and that the deficiencies
      uncovered by the QC system are corrected.   After environmental data are
      collected, QA activities focus on assessing the quality of data obtained
      to determine its suitability to support  enforcement or remedial
      decisions.

1.3   This exhibit describes the overall quality assurance/quality control
      operations and the processes by which the CLP meets the QA/QC objectives
      defined above.  This contract requires a variety of QA/QC activities.
      These contract requirements are the minimum QC  operations necessary to
      satisfy the analytical requirements associated  with the determination of
      the different method analytes.   These QC operations are designed to
      facilitate laboratory comparison by providing EPA with comparable data
      from all Contractors.  These requirements do not release the analytical
      Contractor from maintaining their own QC checks on method and instrument
      performance.
                                     E-4                               OLM03.0

-------
                                                        Exhibit E — Section 2
                                                                  Introduction
2.0   INTRODUCTION

2.1   Appropriate use of data generated under the large range of analytical
      conditions encountered in environmental analyses requires reliance on
      the QC procedures and criteria incorporated into the methods.  The
      ••nethods in this contract have been validated on samples typical of those
      received by the laboratories in the CLP.  However, the validation of
      these methods does not guarantee that they perform equally well for all
      sample matrices encountered.  Inaccuracies can also result from causes
      other than unanticipated matrix effects, such as sampling artifacts,
      equipment malfunctions, and operator error.  Therefore, the quality
      control component of each method is indispensable.

2.2   The data acquired from QC procedures are used to estimate and evaluate
      the information content of analytical results and to determine the
      necessity for or the effect of corrective action procedures.  The
      parameters used to estimate information content include precision,
      accuracy, detection limit, and other quantitative and qualitative
      indicators.  In addition, QC procedures give an overview of the
      activities required in an integrated program to generate data of known
      and documented quality required to meet defined objectives.

2.3   The necessary components of a complete QA/QC program include internal QC
      criteria that demonstrate acceptable levels of performance, as
      determined by QA review.  External review of data and procedures is
      accomplished by the monitoring activities of the National Program
      Of-ice, Regional data users, Sample Management Oifice (SMO), and
      Environme-.tal Monitoring Systems Laboratory/Las Vegas (EMSL/LV).  Each
      external review accomplishes a different purpose.  These reviews are
      described in specific sections of this exhibit.  Laboratory evaluation
      samples, GC/MS tape audits, and data packages provide an external QA
      reference for the program.  A Contractor on-site evaluation system is
      also part of the external QA monitoring.  A feedback loop provides the
      results of the various review functions to the Contractors through
      direct communications with the Technical Project Officers (TPOs) and
      Administrative Project Officers (APOs).

2.4   This exhibit does not provide specific instructions for constructing QA
      plans, QC systems, or a QA organization.  It is, however, an explanation
      of the QA/QC requirements of the program.  It outlines some minimum
      standards for QA/QC programs.   It also includes specific items that are
      required in a QA plan and by the QA/QC documentation detailed in this
      contract.  Delivery of this documentation provides the Agency with a
      complete data package which will stand alone,  and limits the need for
      contact with the Contractor or with an analyst, at a later date, if some
      aspect of the analysis is questioned.

2.5   In order to assure that the product delivered by the Contractor meets
      the requirements of the contract,  and to improve interlaboratory data
      comparison, the Agency requires the following from the Contractor:

      •   Preparation of and adherence to a written quality assurance plan,
          the elements of which are designated in Section 3,
                                      E-5                               OLM03.0

-------
Exhibit E — Section 2
Introduction
      •   Preparation of and adherence to QA/QC standard operating procedures
          as described in Section 4,

      •   Adherence to the analytical methods and associated QC requirements
          specified in the contract,

      •   Verification of analytical standards and documentation of the purity
          of neat materials and the purity and accuracy of solutions obtained
          from private chemical supply houses,

      •   Submission of all raw data and pertinent documentation for Regional
          review,

      •   Participation in the analysis of laboratory evaluation samples,
          including adherence to corrective action procedures,

      •   Submission, upon request, of GC/MS tapes and applicable
          documentation for tape audits, including a copy of the sample data
          package,

      •   Participation in on-site laboratory evaluations, including adherence
          to corrective action procedures, and

      •   Submission of all original documentation generated during sample
          analyses for Agency review.
                                      E-6                               OLM03.0

-------
                                                        Exhibit E — Section 3
                                                        Quality Assurance Plan
3.0   QUALITY ASSURANCE PLAN

3.1   Introduction.  The Contractor shall establish a quality assurance
      program with the objective of providing sound analytical chemical
      measurements.  This program shall incorporate the quality control
      procedures, any necessary corrective action, and all documentation
      required during data collection as well as the quality assessment
      measures performed by management to ensure acceptable data production.

3.1.1    As evidence of such a program, the Contractor shall prepare a written
         quality assurance plan (QAP) which describes the procedures that are
         implemented to achieve the following:

         •   Maintain data integrity, validity, and usability,

         •   Ensure that analytical measurement systems are maintained in an
             acceptable state of stability and reproducibility,

         •   Detect problems through data assessment and establish corrective
             action procedures which keep the analytical process reliable, and

         •   Document all aspects of the measurement process in order to
             provide data which are technically sound and legally defensible.

3.1.2    The QAP shall present, in specific terms, the policies, organization,
         objectives, functional guidelines, and specific QA and QC activities
         designed to achieve the data quality requirements in this contract.
         Where applicable,  standard operating procedures pertaining to each
         element shall be included or referenced as part of the QAP.  The QAP
         shall be paginated consecutively in ascending order.  The QAP shall
         be available during on-site laboratory evaluations.  Additional
         information relevant to the preparation of a QAP can be found in
         Agency and American Society for Testing and Materials publications.

3.2   Required Elements of  a Quality Assurance Plan.  The required elements of
      a laboratory's QAP are outlined in this section.  This outline should be
      used as a framework for developing the QAP.

      A.  Organization and  Personnel

          1.  QA Policy and Objectives

          2.  QA Management

              a.  Organization

              b.  Assignment of QC and QA Responsibilities

              c.  Reporting Relationships

              d.  QA Document Control Procedures

              e.  QA Program Assessment Procedures


                                     E-7                               OLM03.0

-------
Exhibit E — Section 3
Quality Assurance Plan
          3.  Personnel

              a.  Resumes

              b.  Education and Experience

              c.  Training Progress

      B.  Facilities and Equipment

          1.  Instrumentation and Backup Alternatives

          2.  Maintenance Activities and Schedules

      C.  Document Control

          1.  Contractor Notebook Policy

          2.  Sample Tracking/Custody Procedures

          3.  Logbook Maintenance and Archiving Procedures

          4.  Case File Organization,  Preparation and Review Procedures

          5.  Procedures for Preparation,  Approval,  Review,  Revision,  and
              Distribution of Standard Operating Procedures

          6.  Process for Revision of Technical or Documentation Procedures

      D.  Analytical Methodology

          1.  Calibration Procedures and Frequency

          2.  Sample Preparation/Extraction Procedures

          3.  Sample Analysis Procedures

          4.  Standards Preparation Procedures

          5.  Decision Processes,  Procedures, and Responsibility for
              Initiation of Corrective Action

      E.   Data Generation

          I.  Data Collection Procedures

          2.  Data Reduction Procedures

          3.  Data Validation Procedures

          4.  Data Reporting and Authorization  Procedures
                                     E-8                               OLM03.0

-------
                                                        Exhibit E — Section 3
                                                        Quality Assurance Plan
      F.  Quality Control

          1.  Solvent, Reagent and Adsorbent Check Analysis

          2.  Reference Material Analysis

          3.  Internal Quality Control Checks

          4.  Corrective Action and Determination of QC Limit Procedures

          5.  Responsibility Designation

      G.  Quality Assurance

          1.  Data Quality Assurance

          2.  Systems/Internal Audits

          3.  Performance/External Audits

          4.  Corrective Action Procedures

          5.  Quality Assurance Reporting Procedures

          6.  Responsibility Designation

3.3   Updating and Submitting the Quality Assurance Plan

3.3.1    Initial Submission.  During the contract solicitation process, the
         Contractor is required to submit their QAP to the Administrative
         Project Officer.  Within 60 days after contract award,  the Contractor
         shall revise the QAP to be in full compliance with the requirements
         of this contract.  The Contractor shall maintain the QAP on file at
         the Contractor's facility for the term of the contract.  The revised
         QAP will become the official QAP under the contract and may be used
         during legal proceedings.  Both the initial QAP submission and the
         revised QAP shall be paginated consecutively in ascending order.  The
         revised QAP shall include:

         •   Changes resulting from (1) the Contractor's internal review of
             their organization,  personnel, facility, equipment, policy and
             procedures and (2) the Contractor's implementation of the
             requirements of the contract, and

         •   Changes resulting from the Agency's review of the laboratory
             evaluation sample data, bidder-supplied documentation, and
             recommendations made during the pre-award on-site laboratory
             evaluation.

3.3.1.1     The Contractor shall send a copy of the current QAP within 7 days
            of a request from a Technical Project Officer or Administrative
            Project Officer.  The Agency requestor will designate the
            recipients.


                                      E-9                               OLM03.0

-------
Exhibit E — Section 3
Quality Assurance Plan
3.3.2    Subsequent Updates and Submissions.  During the term of the contract,
         the Contractor shall amend the QAP when the following circumstances
         occur:

         •   The Agency modifies the contract,

         •   The Agency notifies the Contractor of deficiencies in the QAP,
                                        t
         •   The Agency notifies the Contractor of deficiencies resulting from
             the Agency's review of the Contractor's performance,

         •   The Contractor identifies deficiencies resulting from the
             internal review of the QAP,

         •   The Contractor's organization, personnel,  facility, equipment,
             policy or procedures change,  or

         •   The Contractor identifies deficiencies resulting from the
             internal review of changes in their organization, personnel,
             facility, equipment, policy or procedures.

3.3.2.1     The Contractor shall amend the QAP within 30 days of when the
            circumstances listed above result in a discrepancy between what
            was previously described in the QAP and what is presently
            occurring at the Contractor's facility.  When the QAP is amended,
            all changes in the QAP shall be clearly marked (e.g., a bar in the
            marain indicating where the change is found in the document, or
            higi.sighting the change by underlining the change, bold printing
            the change, or using a different print font).  The amended pages
            shall have the date on which the changes were implemented.  The
            Contractor shall incorporate all amendments to the current QAP.
            The Contractor shall archive all amendments to the QAP for future
            reference by the Agency.

3.3.2.2     The Contractor shall send a copy of the current QAP within 7 days
            of a request from a Technical Project Officer or Administrative
            Project Officer.  The Agency requestor will designate the
            recipients.

3.4   Corrective Actions.  If the Contractor fails to adhere to the
      requirements listed in Section 3, the Contractor may expect, but the
      Agency is not limited to, the following actions:   reduction of numbers
      of samples sent under this contract, suspension of sample shipment to
      the Contractor, a GC/MS tape audit,  a data package audit, an on-site
      laboratory evaluation, a remedial laboratory evaluation sample, and/or
      contract sanctions, such as a Cure Notice.
                                     E-10                              OLM03.0

-------
                                                        Exhibit E — Section 4
                                                 Standard Operating Procedures
4.0   STANDARD OPERATING PROCEDURES

4.1   Introduction.  In order to obtain reliable results, adherence to
      prescribed analytical methodology is imperative.  In any operation that
      is performed on a repetitive basis, reproducibility is best accomplished
      through the use of standard operating procedures (SOPs).  As defined by
      EPA, an SOP is a written document which provides directions for the
      step-by-step execution of an operation, analysis, or action which is
      commonly accepted as the method for performing certain routine or
      repetitive tasks.

4.1.1    SOPs prepared by the Contractor shall be functional (i.e.,  clear,
         comprehensive, up-to-date, and sufficiently detailed to permit
         duplication of results by qualified analysts).  The SOPs shall be
         paginated consecutively in ascending order.

4.1.2    All SOPs shall reflect activities as they are currently performed by
         the Contractor.   In addition,  all SOPs shall be:

         •   Consistent with current Agency regulations,  guidelines, and the
             CLP contract's requirements.

         •   Consistent with instrument manufacturers' specific instruction
             manuals.

         •   Available to the Agency during an on-site laboratory evaluation.
             A complete set of SOPs shall be bound together and available for
             inspection at such evaluations.   During on-site evaluations,
             Contractor personnel may be asked to demonstrate the application
             of the SOPs.

         •   Available to the designated recipients within 7 days,  upon
             request by the Technical Project Officer or  Administrative
             Project Officer.

         •   Capable of providing for the development of  documentation that is
             sufficiently complete to record the performance of all  tasks
             required by the protocol.

         •   Capable of demonstrating the validity of data reporteU by the
             Contractor and explaining the cause of missing or inconsistent
             results.

         •   Capable of describing the corrective measures and feedback
             mechanism utilized when analytical results do not meet  protocol
             requirements.

         •   Reviewed regularly and updated as necessary  when contract,
             facility,  or Contractor procedural modifications are made.

         •   Archived for future reference in usability or evidentiary
             situations.

         •   Available at specific work stations as appropriate.

                                     E-ll                              OLM03.0

-------
 Exhibit  E  —  Section 4
 Standard Operating Procedures
          •   Subject  to a document  control  procedure  which  precludes  the use
              of  outdated or  inappropriate SOPs.

 4.2    Format.  The  format for  SOPs  may vary depending upon  the  kind of
       activity for  which they  are prepared; however,  at  a minimum, the
       following  sections shall be included:

       •    Title  page,

       •    Scope  and application,

       •    Definitions,

       •    Procedures,

       •    QC  limits,

       •    Corrective  action  procedures, including procedures for secondary
           review of information being generated,

       •    Documentation  description and example forms,

       •    Miscellaneous  notes and precautions, and

       •    References.

4.3   Requirements.  The  Contractor shall maintain the following SOPs.

4.3.1    Evidentiary SOPs  for required chain-of-custody and document  control
         are discussed in  Exhibit F.

4.3.2    Sample  Receipt  and  Storage

         •   Sample receipt  and identification logbooks

         •   Refrigerator  temperature logbooks

         •   Extract storage logbooks

         •   Security precautions

4.3.3    Sample  Preparation

         •   Reagent purity  check procedures and documentation

         •   Extraction procedures

         •   Extraction bench sheets

         •   Extraction  logbook maintenance

4.3.4    Glassware Cleaning



                                     E-12                              OLM03.0

-------
                                                         Exhibit E — Section 4
                                                  Standard Operating Procedures
4.3.5    Calibration  (Balances)

         •   Procedures

         •   Frequency requirements

         •   Preventative maintenance schedule and procedures

         •   Acceptance  criteria and corrective actions

         •   Logbook maintenance

4.3.6    Analytical Procedures  (for each Analytical System, including GPC)

         •   Instrument  performance specifications

         •   Instrumental operating procedures

         •   Data acquisition system operation

         •   Procedures  when automatic  quantitation algorithms are overridden

         •   QC required parameters

         •   Analytical  run/injection logbooks

         •   Instrumental error  and editing flag descriptions and resulting
             corrective  actions

4.3.7    Maintenance Activities  (-for each Analytical System, including GPC)

         •   Preventative maintenance schedule and procedures

         •   Corrective  maintenance determinants and procedures

         •   Maintenance authorization

4.3.8    Analytical Standards

         •   Standard coding/identification and inventory system

         •   Standards preparation logbook(s)

         •   Standards preparation procedures

         •   Procedures  for equivalency/traceability analyses and
             documentation

         •   Purity logbook  (primary standards and solvents)

         •   Storage, replacement,  and  labelling requirements

         •   QC and corrective action measures


                                      E-13                               OLM03.0

-------
Exhibit E — Section 4
Standard Operating Procedures
4.3.9    Data Reduction Procedures

         •   Data processing systems operation

         •   Outlier identification methods

         •   Identification of data requiring corrective action

         •   Procedures for format and/or forms for each operation

4.3.10   Documentation Policy/Procedures

         •   Contractor/analysts' notebook policy, including review policy

         •   Complete SDG File contents

         •   Complete SDG File organization and assembly procedures, including
             review policy

         •   Document inventory procedures, including review policy

4.3.11   Data Validation/Self-Inspection Procedures

         •   Data flow and chain-of-command for data review

         •   Procedures for measuring precision and accuracy

         •   Evaluation parameters for identifying systematic errors

         •   Procedures to ensure that hardcopy and diskette deliverables are
             complete and compliant with the requirements in Exhibits B and H

         •   Procedures to ensure that hardcopy deliverables are in agreement
             with their comparable diskette deliverables

         •   Demonstration of internal QA inspection procedure (demonstrated
             by supervisory sign-off on personal notebooks, internal
             performance evaluation samples, etc.)

         •   Frequency and type of internal audits (e.g., random, quarterly,
             spot checks, perceived trouble areas)

         •   Demonstration of problem identification, corrective actions and
             resumption of analytical processing; sequence resulting from
             internal audit (i.e., QA feedback)

         •   Documentation of audit reports (internal and external), audit
             response,  corrective action, etc.

4.3.12   Data Management and Handling

         •   Procedures for controlling and estimating data entry errors



                                     E-14                              OLM03.0

-------
                                                        Exhibit E — Section 4
                                                 Standard Operating Procedures
         •   Procedures for reviewing changes to data and deliverables and
             ensuring traceability of updates

         •   Life cycle management procedures for testing, modifying and
             implementing changes to existing computing systems including
             hardware, software, and documentation or installing new systems

         •   Database security, backup and archival procedures including
             recovery from system failures

         •   System maintenance procedures and response time

         •   Individuals(s) responsible for system operation, maintenance,
             data integrity and security

         •   Specifications for staff training procedures

         •   Storage, retrieval and verification of the completeness and
             readability of GC/MS files transferred to magnetic media

4.4   Submitting and Updating SOPs

4.4.1    Initial Submission.   During the contract solicitation process, the
         Contractor is required to submit their SOPs to the Administrative
         Project Officer (APO).  Within 60 days after contract award, the
         Contractor shall prepare and maintain on file, at their facility, a
         complete,  revised set of SOPs fully compliant with the requirements
         of this contract.   The revised SOPs will become the official SOPs
         under the contract and may be used during legal proceedings.  Both
         the initial submission of SOPs and the revised SOPs shall be
         paginated consecutively in ascending order.  The revised SOPs shall
         include:

         •   Changes resulting from (1) the Contractor's internal review of
             their procedures and (2)  the Contractor's implementation of the
             requirements of the contract, and

         •   Changes resulting from the Agency's review of the laboratory
             evaluation sample data, bidder-supplied documentation, and
             recommendations made during the pre-award on-site laboratory
             evaluation.

4.4.1.1   The Contractor shall send a complete set of current SOPs or
          individually requested SOPs within 7 days of a request from a
          Technical Project Officer or Administrative Project Officer.  The
          Agency requestor will designate the recipients.

4.4.2    Subsequent Updates and Submissions.  During the term of the contract,
         the Contractor shall amend the SOPs when the following circumstances
         occur:

         •   The Agency modifies the contract,

         •   The Agency notifies the Contractor of deficiencies in their SOPs,

                                     E-15                              OLM03.0

-------
 Exhibit  E  —  Section  4
 Standard Operating  Procedures


         •   The  Agency  notifies the Contractor of deficiencies resulting  from
              the  Agency's  review of the Contractor's performance,

         •   The  Contractor's procedures change,

         f   The  Contractor  identifies deficiencies resulting from the
              internal review of their SOPs documentation, or

         •   The  Contractor  identifies deficiencies resulting from the
              internal review of their procedures.

 4.4.2.1     Existing SOPs  shall be amended or new SOPs shall be written within
            30 days of when  the circumstances listed above result in a
            discrepancy  between what was previously described in the SOPs  and
            what  is presently occurring at the Contractor's facility.  All
            changes in the SOPs shall be clearly marked  (e.g., a bar in the
            margin  indicating where the change is in the document, or
            highlighting the change by underlining the change, bold printing
            the change,  or using a different print font).  The amended/new
            SOPs  shall have  the date on which the changes were implemented.

 4.4.2.2     When  existing SOPs are amended or new SOPs are written, the
            Contractor shall document the reason(s) for the change, and
            maintain the amended or new SOPs on file at the laboratory
            facility.  Documentation of the reason(s)  for the changes shall be
            maintained on file with the amended SOPs or new SOPs.

 4.4.2.3     The Contractor shall send a complete set of current SOPs or
            individually requested SOPs within 7 days of a request from a
            Technical Project Officer or Administrative Project Officer.  The
            Agency requestor will designate the recipients.

 4.4.2.4     Documentation of the reason(s)  for changes to the SOPs shall also
            be submitted with the SOPs.   An alternate delivery schedule for
            submitting the amended/new SOPs and their documentation  may be
            proposed by the Contractor,  but it is the sole decision of the
            Agency,  represented either by the Technical Project Officer or
            Administrative Project Officer, to approve or disapprove the
            alternate delivery schedule.   If an alternate delivery schedule is
            proposed,  the Contractor shall  describe in a letter to the
            Technical Project Officer,  Administrative Project Officer, and the
            Contracting Officer why he/she  is unable to meet the delivery
            schedule listed in this section.   The Technical Project
            Officer/Administrative Project  Officer will not grant an extension
            for greater than 30 days for amending/writing new SOPs.  The
            Technical Project Officer/Administrative Project Officer will not
            grant an extension for greater  than 14 days for submission of the
            letter documenting the reasons  for the changes and for submitting
            amended/new SOPs.  The Contractor shall proceed and not assume
            that an extension will be granted until so notified by the
            Technical Project Officer and/or Administrative Project Officer.

4.5   Corrective Actions.  If the Contractor fails to adhere to the
      requirements listed in Section 4,  the Contractor may expect, but the

                                     E-16                              OLM03.0

-------
                                                  Exhibit E — Section 4
                                           Standard Operating Procedures
Agency is not limited to, the following action:  reduction of number of
samples sent under this contract, suspension of sample shipment to the
Contractor, a GC/MS tape audit, a data package audit, an on-site
laboratory evaluation, a remedial laboratory evaluation sample, and/or
contract sanctions, such as a Cure Notice.
                               E-17                              OLM03.0

-------
Exhibit E — Section 5
Analytical Standards Requirements
5.0   ANALYTICAL STANDARDS REQUIREMENTS

5.1   Overview.  EPA will not supply analytical reference standards either for
      direct analytical measurements or for the purpose of traceability.  All
      Contractors shall be required to prepare from neat materials or purchase
      from private chemical supply houses those standards necessary to
      successfully and accurately perform the analyses required in this
      protocol.

5.2   Preparation of Chemical Standards from the Neat High Purity Bulk
      Material.  A Contractor may prepare their chemical standards from neat
      materials.  Contractors shall obtain the highest purity possible when
      purchasing neat chemical standards; when standards are purchased at less
      than 97% purity, the Contractor shall document the reason why a higher
      purity could not be obtained.

5.2.1    Neat chemical standards shall be kept refrigerated when not being
         used in the preparation of standard solutions.  Proper storage of
         neat chemicals is essential in order to safeguard them from
         decomposition.

5.2.2    The purity of a compound can sometimes be misrepresented by a
         chemical supply house.   Since knowledge of purity is needed to
         calculate the concentration of solute in a solution standard, it is
         the Contractor's responsibility to have analytical documentation
         ascertaining that the purity of each compound is correctly stated.
         Purity confirmation,  when performed, should use either differential
         scanning calorimetry, gas chromatography with flame ionization
         detection,  high performance liquid chromatography, infrared
         spectrometry, or other appropriate techniques.  Use of two or more
         independent methods is recommended.  The correction factor for
         impurity when weighing neat materials in the preparation of solution
         standards is:
         EQ. 1
              weight of impure compound = ^ight of pure compound
                                           (percent purity/100)
         where "weight of pure compound" is that required to prepare a
         specific volume of a standard solution at a specified concentration,

5.2.3    When compound purity is assayed to be 97% or greater, the weight may
         be used without correction to calculate the concentration of the
         stock standard.  If the compound purity is assayed to be less than
         97%, the weight shall be corrected when calculating the concentration
         of the stock solution.
                                     E-18                              OLM03.0

-------
                                                        Exhibit E — Section 5
                                             Analytical Standards Requirements
5.2.4    Mis-identification of compounds occasionally occurs and it is
         possible that a mislabeled compound may be received from a chemical
         supply house.  It is the Contractor's responsibility to have
         analytical documentation ascertaining that all compounds used in the
         preparation of solution standards are correctly identified.
         Identification confirmation, when performed, shall use gas
         chromatography/mass spectrometry analysis on at least two different
         analytical columns, or other appropriate techniques.

5.2.5    Calculate the weight of material to be weighed out for a specified
         volume taking into account the purity of the compound and the desired
         concentration.  A second person shall verify the accuracy of the
         calculations.  Check balances for accuracy with a set of standard
         weights every 12 hours.  All weighing shall be performed on an
         analytical balance to the nearest 0.1 mg and verified by a second
         person.  The solvent used to dissolve the solute shall be compatible
         with the protocol in which the standard is to be used; the solute
         shall be soluble, stable,  and nonreactive with the solvent.  In the
         case of a multicomponent solution, the components must not react with
         each other.

5.2.6    Transfer the solute to a volumetric flask and dilute to the specified
         solution volume with solvent after ensuring dissolution of the solute
         in the solvent.  Sonication or warming may be performed to promote
         dissolution of the solute.  This solution shall be called the primary
         standard and all subsequent dilutions shall be traceable back to the
         primary standard.

5.2.7    Log notebooks shall be kept for all weighing and dilutions.  All
         subsequent dilutions from the primary standard and the calculations
         for determining their concentrations shall be recorded and verified
         by a second person.   All solution standards shall be refrigerated
         when not in use.   All solution standards shall be clearly labeled as
         to the identity of the compound or compounds, concentration,  date
         prepared, solvent, and initials of the preparer.

5.3   Purchase of Chemical Standards Already in Solution.  Solutions of
      analytical reference standards can be purchased by Contractors provided
      they meet the following criteria.

5.3.1    Contractors shall maintain the following documentation to verify the
         integrity of the standard solutions they purchase:

         •   Mass spectral identification confirmation of the solution,

         •   Purity confirmation of the solution, and

         •   Chromatographic and quantitative documentation that the solution
             standard was QC checked according to the following section.

5.3.2    The Contractor shall purchase standards for which the quality is
         demonstrated statistically and analytically.   One way this may be
         demonstrated is to prepare and analyze three solutions, a high
         standard, a low standard,  and a standard at the target concentration

                                     E-19                              OLM03.0

-------
Exhibit E — Section 5
Analytical Standards Requirements
         (see Sections 5.3.2.1 and 5.3.2.2).  The Contractor shall have
         documentation to demonstrate that the analytical results for the high
         standard and low standard are consistent with the difference in
         theoretical concentrations.   This is done by the Student's t-test in
         Section 5.3.2.4.  If this is achieved,  the Contractor shall then
         demonstrate that the concentration of the target standard lies midway
         between the concentrations of the low and high standards.  This is
         done by the Student's t-test in Section 5.3.2.5.  The standard is
         certified to be within 10% of the target concentration using the
         equations in Section 5.3.2.6.  If this procedure is used, the
         Contractor shall document that the following have been achieved.

5.3.2.1     Two solutions of identical concentration shall be prepared
            independently from solutions.   An aliquot of the first solution
            shall be diluted to the intended concentration (the "target
            standard").   One aliquot  is taken from the second solution and
            diluted to a concentration 10% greater than the target standard.
            This is called the "high  standard."   One further aliquot is taken
            from the second solution  and diluted to a concentration 10% less
            that the target standard.  This is called the "low standard."

5.3.2.2     Six replicate analyses of each standard (a total of 18 analyses)
            shall be performed in the following sequence: low standard,
            target, high standard, low standard, target standard,  high
            standard,  ...

5.3.2.3     The mean and variance of  the six results for each solution shall
            be calculated.

            EQ. 2
                                 Mean = -
            EQ.  3

                                      6
                                     £ Yl - 6 (MEAN) z
                          Variance = —	
                                     E-20                              OLM03.0

-------
                                                         Exhibit  E  —  Section  5
                                             Analytical  Standards  Requirements
            The values  Yj represent the results of the six analyses of each
            standard.   The means  of the low, target,  and  high  standards  are
            designated  M^, M2, and Mj,  respectively.   The  variances of the
            low, target, and  high standards are designated Vi, Vj, and V},
            respectively.  Additionally,  a pooled variance, Vp, is calculated.
            EQ.  4
                                               V,
                                  0.81
                                  1.21
If the  square root of Vp is less than 1% of
                                                            then M2 /10,000
            shall be used as the value of Vp in all subsequent calculations.

5.3.2.4     The test statistic shall be calculated.

            EQ. 5
                          Test Statistic =
                                           1.1
                                                 0.9
5.3.2.5
If une test statistic exceeds 2.13 then a 20% difference between
the high and low standards exists.  In such a case, the standards
are not acceptable.

The test statistic shall be calculated.

EQ. 6
                        Test Statistic =
                                             1.8
                                       2.2
5.3.2.6
If the test statistic exceeds 2.13, then the target standard
concentration has not been demonstrated to be the midway between
the high and low standards.  In such a case, the standards are not
acceptable.

The 95% confidence intervals for the mean result of each standard
shall be calculated.
                                     E-21
                                                           OLM03.0

-------
Exhibit E — Section 5
Analytical Standards Requirements
            EQ. 7


                   Interval for Low Standard = M^ ± 2 .13



            EQ. 8


                  Interval for Target Standard = M2 ± 2.13



            EQ. 9


                   Interval for High Standard = M3 ± 2.13
5.3.2.6.1      These intervals shall not overlap.   If overlap is observed, the
               ability to discriminate the 10% difference in concentrations
               has not been demonstrated.  In such a case,  the standards are
               not acceptable.

5.3.2.6.2      In any event, the Contractor is responsible for the quality of
               the standards employed for analyses under this contract.

5.4   Requesting Standards From the EPA Standards  Repository.  Solutions of
      analytical reference materials can be ordered from the U.S.  EPA Chemical
      Standards Repository, depending on availability.   The Contractor can
      place an order for standards only after demonstrating that these
      standards are not available from commercial  vendors,  either in solution
      or as a neat material.

5.5   Documentation of the Verification and Preparation of Chemical Standards.
      It is the responsibility of each Contractor  to maintain the necessary
      documentation to show that the chemical standards they have used in the
      performance of CLP analysis conform to the requirements previously
      listed.

5.5.1    Weighing logbooks, calculations, chromatograms, mass spectra, etc.,
         whether produced by the Contractor or purchased from chemical supply
         houses, shall be maintained by the Contractor and may be subject to
         review during on-site laboratory evaluations.   In those cases where
         the documentation is supportive of the analytical results of data
         packages sent to the Agency, such documentation is to be kept on file
         by the Contractor for a period of one year.

5.5.2    Upon request by the Technical Project Officer or Administrative
         Project Officer, the Contractor shall submit their most recent
         previous year's documentation (12 months) for the verification and
         preparation of chemical standards within 14 days of the receipt of
         request to the recipients he/she designates.
                                     E-22                              OLM03.0

-------
                                                        Exhibit E — Section 5
                                             Analytical Standards Requirements
5.5,3    The Agency may generate a report discussing deficiencies in the
         Contractor's documentation for the verification and preparation of
         chemical standards or may discuss the deficiencies during an on-site
         laboratory evaluation.  In a detailed letter to the Technical Project
         Officer, Administrative Project Officer, and EMSL/LV, the Contractor
         shall address the deficiencies and the subsequent corrective action
         implemented by the Contractor to correct the deficiencies within 14
         days of receipt of the report or the on-site laboratory evaluation.
         An alternate delivery schedule may be proposed by the Contractor, but
         it is the sole decision of the Agency, represented either by the
         Technical Project Officer or Administrative Project Officer, to
         approve or disapprove the alternate delivery schedule.  If an
         alternate delivery schedule is proposed, the Contractor shall
         describe in a letter to the Technical Project Officer, Administrative
         Project Officer, and the Contracting Officer why he/she is unable to
         meet the delivery schedule listed in this section.  The Technical
         Project Officer/Administrative Project Officer will not grant an
         extension for greater than 14 days for the Contractor's response
         letter to the standards documentation report.  The Contractor shall
         proceed and not assume that an extension will be granted until so
         notified by the Technical Project Officer and/or Administrative
         Project Officer.

5.5.4    If new SOPs are required to be written or SOPs are required to be
         amended because of deficiencies and the subsequent corrective action
         implemented by the Contractor, the Contractor shall write/amend and
         submit the SOPs per the requirements listed in Section 4.

5.6   Corrective Actions.  If the Contractor fails to adhere to the
      requirements listed in Section 5, the Contractor may expect, but the
      Agency is not limited to, the following actions:  reduction of number of
      samples sent under the contract, suspension of sample shipment to
      Contractor, a GC/MS tape audit,  a data package audit, an on-site
      laboratory evaluation, a remedial laboratory evaluation sample, and/or
      contract sanctions, such as a Cure Notice.
                                     E-23                              OLM03.0

-------
Exhibit E — Section 6
Contract Compliance Screening
6.0   CONTRACT COMPLIANCE SCREENING

6.1   Contract compliance screening (CCS)  is one aspect of the Government's
      contractual right of inspection of analytical data.  CCS examines the
      Contractor's adherence to the contract requirements  based on the sample
      data package delivered to the Agency.

6.2   CCS is performed by the Sample Management Office (SMO)  under the
      direction of the Agency.   To assure a  uniform review,  a set of
      standardized procedures has been developed to evaluate the sample data
      package submitted by a Contractor against the technical and completeness
      requirements of the contract.  The government reserves the right to add
      and/or delete individual checks.  CCS  results are used in conjunction
      with other information to measure overall Contractor performance and to
      take appropriate actions to correct deficiencies in  performance.

6.3   CCS results are mailed to the Contractor and all other data recipients.
      The Contractor has a period of time to correct deficiencies.  The
      Contractor shall send all corrections  to the Regional client and SMO.

6.4   The Agency may generate a CCS trend report which summarizes CCS results
      over a given period of time.   The Agency may send the CCS trend report
      or discuss the CCS trend report during an on-site laboratory evaluation.
      In a detailed letter to the Technical  Project Officer and Administrative
      Project Officer, the Contractor shall  address the deficiencies and the
      subsequent corrective action implemented by the Contractor to correct
      the deficiencies within 14 days of receipt of the report or the on-site
      laboratory evaluation.  An alternate delivery schedule may be proposed
      by the Contractor, but it is the sole  decision of the Agency,
      represented by the Technical Project Officer or Administrative Project
      Officer, to approve or disapprove the  alternate delivery schedule.  If
      an alternate delivery schedule is proposed, the Contractor shall
      describe in a letter to the Technical  Project Officer,  Administrative
      Project Officer, and Contracting Officer why he/she  is unable to meet
      the delivery schedule listed in this section.  The Technical Project
      Officer/Administrative Project Officer will not grant an extension for
      greater than 14 days for the Contractor's response to the CCS trend
      report.  The Contractor shall proceed  and not assume that an extension
      will be granted until so notified by the Technical Project Officer
      and/or Administrative Project Officer.

6.5   If new SOPs are required to be written or SOPs are required to be
      amended because of the deficiencies and the subsequent corrective action
      implemented by the Contractor, the Contractor shall  write/amend and
      submit the SOPs per the requirements listed in Section 4.

6.6   If the Contractor fails to adhere to the requirements listed in Section
      6, the Contractor may expect, but the  Agency is not  limited to, the
      following actions:  reduction of number of samples sent under the
      contract, suspension of sample shipment to the Contractor, a GC/MS tape
      audit, a data package audit, an on-site laboratory evaluation, a
      remedial laboratory evaluation sample, and/or contract sanctions, such
      as a Cure Notice.
                                     E-24                              OLM03.0

-------
                                                        Exhibit E — Section 7
                                                          Regional Data Review
7.0   REGIONAL DATA REVIEW

7.1   Contractor data are generated to meet the specific needs of the EPA
      Regions. In order to verify the usability of data for the intended
      purpose, each Region reviews data from the perspective of the end user,
      based upon functional aspects of data quality.  General guidelines for
      data review have been developed jointly by the Regions and the National
      Program Office.  Each Region uses these guidelines as the basis for data
      evaluation.  Individual Regions may augment the basic guideline review
      process with additional review based on Region-specific or site-specific
      concerns.  Regional reviews, like the sites under investigation, vary
      based on the nature of the problems under investigation and the Regional
      response appropriate to the specific circumstances.

7.2   Regional data reviews, relating usability of the data to a specific
      site, are part of the collective assessment process.   They complement
      the review done at the Sample Management Office, which is designed to
      identify contractual discrepancies,  and the review done at EMSL/LV,
      which is designed to evaluate Contractor and method performance.  These
      individual evaluations are integrated into a collective review that is
      necessary for Program and Contractor administration and management and
      may be used to take appropriate action to correct deficiencies in the
      Contractor's performance.
                                     E-25                              OLM03.0

-------
Exhibit E — Section 8
Laboratory Evaluation Samples
8.0   LABORATORY EVALUATION SAMPLES

8.1   Although intralaboratory QC may demonstrate contractor and method
      performance that may be tracked over time, an external performance
      evaluation program is an essential feature of a QA program.  As a means
      of measuring Contractor and method performance, Contractors participate
      in interlaboratory comparison studies conducted by the Agency.  Results
      from the analysis of these laboratory evaluation samples, also referred
      to as performance evaluation (PE) samples, will be used by the Agency to
      verify the Contractor's continuing ability to produce acceptable
      analytical data. The results are also used to assess the precision and
      accuracy of the analytical methods for specific analytes.

8.2   Sample sets may be provided to participating Contractors as frequently
      as on an SDG-by-SDG basis as a recognizable QC sample of known
      composition, as a recognizable QC sample of unknown composition, or not
      recognizable as a QC material.   The laboratory evaluation samples may be
      sent either by the Regional client or the National Program Office.  The
      results of all such quality control samples may be used as the basis for
      an equitable adjustment to reflect the reduced value to the Agency; may
      be used as the basis for rejection of data for: sample(s) within an SDG,
      a fraction within an SDG or the entire SDG; and/or may be used as the
      basis for contract action.  The Contractor shall analyze the samples and
      return the data package and all raw data within the contract required
      turnaround time.

8.3   At a minir-im, the results are evaluated fo^- compound identification,
      quantitation, and sample contamination.  Confidence intervals for the
      quantitation of target compounds are based on reported values using
      population statistics.  The Agency may adjust the scores on any given
      laboratory evaluation sample to compensate for unanticipated
      difficulties with a particular sample.  Normally, a fraction of the
      compounds spiked into the sample are not specifically listed in the
      contract.  Contractors are required to use the NIST/EPA/NTH (May 1992
      release or later) and/or Wiley (1991 release or later), or equivalent,
      mass spectral library to tentatively identify a maximum number of non-
      target compounds in each fraction that are present above a minimal
      response.  Tentative identification of these compounds, based on
      contractually described spectral interpretation procedures, is evaluated
      and integrated into the evaluation process.

8.4   A Contractor's results on the laboratory evaluation samples will
      determine the Contractor's performance as follows.

8.4.1    Acceptable, No Response Required (Score greater than or equal to
         90%):  Data meets most or all of the scoring criteria.

8.4.2    Acceptable, Response Explaining Deficiency(ies) Required (Score
         greater than or equal to 75% but less than 90%):  Deficiencies exist
         in the Contractor's performance.

8.4.3    Unacceptable Performance, Response Explaining Deficiency(ies)
         Required (Score less than 75%):  Deficiencies exist in the
         Contractor's performance to the extent that the National Program

                                     E-26                              OLM03.0

-------
                                                        Exhibit E — Section 8
                                                 Laboratory Evaluation Samples
         Office has determined that the Contractor has not demonstrated the
         capability to meet the contract requirements.

8.5   In the case of Sections 8.4.2 and 8.4.3, the Contractor shall describe
      the deficiency(ies) and the action(s) taken to correct the
      deficiency(ies) in a letter to the Administrative Project Officer, the
      Technical Project Officer and EMSL/LV within 14 days of receipt of
      notification from the Agency.

8.6   An alternate delivery schedule may be proposed by the Contractor, but it
      is the sole decision of the Agency, represented either by the Technical
      Project Officer or Administrative Project Officer, to approve or
      disapprove the alternate delivery schedule.  If an alternate delivery
      schedule is proposed, the Contractor shall describe in a letter to the
      Technical Project Officer, Administrative Project Officer, and the
      Contracting Officer why he/she is unable to meet the delivery schedule
      listed in this section.  The Technical Project Officer/Administrative
      Project Officer will not grant an extension for greater than 14 days for
      the Contractor's response letter to the laboratory evaluation sample
      report.  The Contractor shall proceed and not assume that an extension
      will be granted until so notified by the Technical Project Officer
      and/or Administrative Project Officer.

8.7   If new SOPs are required to be written or SOPs are required to be
      amended because of the deficiencies and the subsequent corrective action
      implemented by the Contractor, the Contractor shall write/amend and
      submit the SOPs per the requirements listed in Section 4.

8.8   The Contractor shall be notified by the Technical Project Officer,
      Administrative Project Officer,  or Contracting Officer concerning the
      remedy for their unacceptable performance.  The Contractor may expect,
      but the Agency is not limited to, the following actions:  reduction of
      the number of samples sent under the contract, suspension of sample
      shipment to the Contractor,  a GC/MS tape audit, a data package audit, an
      on-site laboratory evaluation, a remedial laboratory evaluation sample,
      and/or contract sanctions, such as a Cure Notice.

      NOTE:  The Contractor's prompt response demonstrating that corrective
      actions have been taken to ensure the Contractor's capability to meet
      contract requirements may facilitate continuation of sample scheduling.

8.9   If the Contractor fails to adhere to the requirements listed in Section
      8, the Contractor may expect, but the Agency is not limited to, the
      following actions:  reduction in the number of samples sent under the
      contract, suspension of sample shipment to the Contractor, a GC/MS tape
      audit, a data package audit,  an on-site laboratory evaluation, a
      remedial laboratory evaluation sample and/or contract sanctions, such as
      a Cure Notice.
                                     E-27                              OLM03.0

-------
Exhibit E — Section 9
GC/MS Tape Audits
9.0   GC/MS TAPE AUDITS

9.1   Overview.  Periodically,  the Agency requests the GC/MS magnetic tapes
      from Contractors for a specific Case in order to accomplish tape audits.
      Generally, tape submissions and audits are requested for the following
      reasons.

      •   Program overview,

      •   Indication of data quality problems from EMSL/LV,  SMO,  or Regional
          data reviews,

      •   Support for on-site audits, and

      •   Specific Regional  requests.

9.1.1    Depending upon the  reason for an audit, the tapes from a recent Case,
         a specific Case,  or a laboratory evaluation sample may be requested.
         Tape audits provide a mechanism to assess adherence to contractual
         requirements and to ensure the consistency of data reported on the
         hardcopy/floppy diskettes with that generated on the GC/MS tapes.
         This function provides external monitoring of Program QC requirements
         and checks adherence of the Contractor to internal QA procedures.  In
         addition, tape audits enable the Agency to evaluate the utility,
         precision, and accuracy of the analytical methods.

9.1.2    The Contractor shall store all raw and processed GC/MS data on
         magnetic tape, in appropriate instrument manufacturer's format.  This
         tape shall include  data for samples, blanks,  matrix spikes, matrix
         spike duplicates, initial calibrations, continuing calibrations, and
         instrument performance checks (BFB and DFTPP) as well as all
         Contractor-generated spectral libraries and quantitation reports
         required to generate the data package.  The Contractor shall maintain
         a written reference logbook of tape files of the EPA sample number,
         calibration data, standards, blanks, matrix spikes, and matrix spike
         duplicates.  The logbook shall include EPA sample numbers and
         standard and blank  Ids, identified by Case and Sample Delivery Group.

9.1.3    The Contractor is required to retain the GC/MS tapes for 365 days
         after submission of the reconciled Complete SDG File.  When
         submitting GC/MS tapes to the Agency, the following materials shall
         be delivered in response to the request.

9.1.3.1     All associated raw data files for samples, including laboratory
            evaluation samples, blanks, matrix spikes, matrix spike
            duplicates, initial and continuing calibration standards, and
            instrument performance check solutions (BFB and DFTPP).

9.1.3.2     All processed data files and quantitation output files associated
            with the raw data files described in Section 9.1.3.1.

9.1.3.3     All associated identifications and calculation files used to
            generate the data submitted in the data package.
                                     E-28                              OLM03.0

-------
                                                        Exhibit E — Section 9
                                                             GC/MS Tape Audits
9.1.3.4     All Contractor-generated mass spectral library files  (NIST/EPA/NIH
            and/or Wiley, or equivalent, library not required).

9.1.3.5     A copy of the Contractor's written reference logbook  relating tape
            files to EPA sample number, calibration data, standards, blanks,
            matrix spikes, and matrix spike duplicates.  The logbook shall
            include EPA sample numbers and lab file identifiers for all
            samples, blanks, and standards, identified by Case and Sample
            Delivery Group.

9.1.3.6     A directory of files on each tape.

9.1.3.7     A copy of the completed sample data package.

9.1.3.8     A statement attesting tc the completeness of the GC/MS data tape
            submission, signed and dated by the Contractor's laboratory
            manager.  The Contractor shall also provide a statement attesting
            that the data reported have not been altered in any way.  These
            statements shall be part of a cover sheet that includes the
            following information relevant to the data tape submission:

            •   Contractor name,

            •   Date of submission,

            •   Case number,

            •   SDG number,

            •   GC/MS make and model number,

            •   Software version,

            •   Disk drive type (e.g.,  CDC,  PRIAM,  etc.),

            •   File transfer method (e.g.,  DSD,  DTD,  FTP,  Aguari'is, etc.),
                and

            •   Names and telephone numbers of two Contractor contacts for
                further information regarding the submission.

9.2   Submission of the GC/MS Tape.  Upon request of the Administrative
      Project Officer or EMSL/LV,  the Contractor shall send the required GC/MS
      tapes and all necessary documentation to EMSL/LV or designated recipient
      within seven (7)  days of notification.   An alternate delivery schedule
      may be proposed by the Contractor,  but  it is the sole decision of the
      Agency,  represented either by the Technical Project Officer or
      Administrative Project Officer,  to approve or disapprove the alternate
      delivery schedule.  If an alternate delivery schedule is proposed, the
      Contractor shall describe in a letter to the Technical Project Officer,
      Administrative Project Officer,  and the Contracting Officer why he/she
      is unable to meet the delivery schedule listed in this section.   The
      Technical Project Officer/Administrative Project Officer will not grant


                                     E-29                              OLM03.0

-------
Exhibit E — Section 9
GC/MS Tape Audits
      an extension for greater than seven days for submission of the GC/MS
      tape.  The Contractor shall proceed and not assume that an extension
      will be granted until so notified by the Technical Project Officer
      and/or Administrative Project Officer.

      NOTE:  The GC/MS tapes shall be shipped according to the procedures in
      Exhibit F.

9.3   Responding to the GC/MS Tape Audit Report.   After completion of the
      GC/MS tape audit, the Agency may send a copy of the GC/MS tape audit
      report to the Contractor or may discuss the GC/MS tape audit report at
      an on-site laboratory evaluation.  In a detailed letter to the Technical
      Project Officer, Administrative Project Officer, and EMSL/LV, the
      Contractor shall discuss the corrective actions implemented to resolve
      the deficiencies listed in the GC/MS tape audit report within 14 days of
      receipt of the report.

9.3.1    An alternate delivery schedule may be proposed by the Contractor, but
         it is the sole decision of the Agency,  represented either by the
         Technical Project Officer or Administrative Project Officer, to
         approve or disapprove the alternate delivery schedule.  If an
         alternate delivery schedule is proposed,  the Contractor shall
         describe in a letter to the Technical Project Officer, Administrative
         Project Officer,  and the Contracting Officer why he/she is unable to
         meet the delivery schedule listed in this section.   The Technical
         Project Officer/Administrative Project Officer will not grant an
         extension for greater than 14 days for the Contractor's response
         letter to the GC/MS tape report.  The Contractor shall proceed and
         not assume that an extension will be granted until so notified by the
         Technical Project Officer and/or Administrative Project Officer.

9.3.2    If new SOPs are required to be written or SOPs are required to be
         amended because of the deficiencies and the subsequent corrective
         action implemented by the Contractor, the Contractor shall
         write/amend and submit the SOPs per the requirements listed in
         Section 4.

9.3.3    Maintenance of the Magnetic Tape Storage Device

9.3.3.1     The Contractor shall certify that the tape head alignment on the
            magnetic tape storage device is in compliance with the ANSI
            standards for nine track magnetic tapes.  If the Contractor does
            not have documentation of alignment within the last 12 months, the
            Contractor must perform or have performed the manufacturer's
            documented head alignment procedure within 60 days of contract
            award.  This is generally performed with a "skew" tape, certified
            to be in conformance with ANSI standards.  The alignment must be
            performed by qualified personnel.  The tape head alignment must be
            performed at a minimum once every 12 months or when there is
            evidence that the tape head may be out of alignment.
                                     E-30                              OLM03.0

-------
                                                        Exhibit E — Section 9
                                                             GC/MS Tape Audits
9.3.3.2     The tape system, including recording head, must be in conformance
            with the manufacturer's physical and electrical standards.
            Alignment of the remaining components of the tape system such as
            the retracting arms, must be performed at intervals not to exceed
            24 months.  If the Contractor cannot demonstrate that the
            remaining components of the t ipe system are in alignment, then the
            Contractor must perform or have performed the manufacturer's
            recommended alignment procedure.

9.3.3.3     Documentation of maintenance, alignment, and repair procedures
            must be kept in an instrument maintenance log book for each tape
            device and data system.  Also include any local area network
            components that provide a means for the transmission of data to or
            from the instrument data system and the tape system.  Maintenance
            entries must include serial number, property number (if
            applicable), data and time of repair, name of person performing
            maintenance, problem description, problem resolution, date and
            time of failure (if applicable), and date and time placed back in
            service.  Copies of repairs shall be kept in the maintenance
            documentation.  Documentation of 1) data system, and 2) tape
            system maintenance and alignments, for the last 24 months must be
            made available upon written request of the TPO, APO or EMSL/LV or
            during a laboratory on-site evaluation.  The Contractor shall
            always submit a GC/MS tape from a tape system in conformance with
            the manufacturer's physical and electrical standards and alignment
            according to manufacturer's procedures.

9.4   Corrective Actions.  If the Contractor fails to adhere to the
      requirements listed in Section 9, the Contractor may expect, but the
      Agency is not limited to,  the following actions:  reduction in the
      number of samples sent under the contract, suspension of sample shipment
      to the Contractor, an on-site laboratory evaluation, a GC/MS tape audit,
      a data package audit, a remedial laboratory evaluation sample, and/or
      contract sanctions, such as a Cure Notice.
                                     E-31                              OLM03.0

-------
Exhibit E — Section 10
Data Package Audits
10.0  DATA PACKAGE AUDITS

10.1  Overview.  Data package audits are performed by the Agency for program
      overview and specific Regional concerns and to assess the technical
      quality of the data and evaluate overall Contractor performance.  They
      provide the Agency with an in-depth inspection and evaluation of the
      Case data package with regard to achieving QA/QC acceptability.  Data
      packages are periodically selected from recently received Cases.  They
      are evaluated for the technical quality of hardcopy raw data, quality
      assurance, and adherence to contractual requirements.  A thorough review
      of the raw data is completed, including:  a check of instrument
      printouts, quantitation reports, chromatograms, spectra, library
      searches and other documentation for deviations from the contractual
      requirements, a check for transcription and calculation errors, a review
      of the qualifications of the Contractor personnel involved with the
      Case,  and a review of all current SOPs on file.  Standardized procedures
      have been established to assure uniformity of the auditing process.

10.2  Responding to the Data Package Audit Report.   After completing the data
      package audit,  the Agency may send a. copy of  the data package audit
      report to the Contractor or may discuss the data package audit report at
      an on-site laboratory evaluation.  In a detailed letter to the Technical
      Project Officer,  Administrative Project Officer, and EMSL/LV, the
      Contractor shall  discuss the corrective actions implemented to resolve
      the deficiencies  listed in the data package audit report within 14 days
      of receipt of the report.

10.2.1   An alternate delivery schedule may be proposed by the Contractor, but
         it is the sole decision of the Agency,  represented either by the
         Technical Project Officer or Administrative Project Officer, to
         approve or disapprove the alternate delivery schedule.  If an
         alternate delivery schedule is proposed, the Contractor shall
         describe in a  letter to the Technical Project Officer, Administrative
         Project Officer,  and the Contracting Officer, why he/she is unable to
         meet the delivery schedule listed in this  section.  The Technical
         Project Officer/Administrative Project Officer will not grant an
         extension for  greater than 14 days for the Contractor's response
         letter to the  data package report.  The Contractor shall proceed and
         not assume that an extension will be granted until so notified by the
         Technical Project Officer and/or Administrative Project Officer.

10.2.2   If new SOPs are required to be written or  SOPs are required to be
         amended because of the deficiencies and the subsequent corrective
         action implemented by the Contractor, the Contractor shall
         write/amend and submit the SOPs per the requirements listed in
         Section 4.

10.3  Corrective Actions.   If the Contractor fails  to adhere to the
      requirements listed in Section 10, the Contractor may expect, but the
      Agency is not limited to, the following actions:  reduction in the
      numbers of samples sent under the contract, suspension of sample
      shipment to the Contractor, an on-site laboratory evaluation, a GC/MS
      tape audit, a data package audit, a remedial laboratory evaluation
      sample, and/or contract sanctions, such as a Cure Notice.

                                     E-32                              OLM03.0

-------
                                                       Exhibit E — Section 11
                                                On-Site Laboratory Evaluations
11.0  ON-SITE LABORATORY EVALUATIONS

11.1  Overview.  At a frequency dictated by a Contractor's performance, the
      Administrative Project Officer, Technical Project Officer or the
      Contracting Officer will conduct an on-site laboratory evaluation.  On-
      site laboratory evaluations are Carried out to monitor the Contractor's
      ability to meet selected terms and conditions specified in the contract.
      The evaluation process incorporates two separate categories:  a quality
      assurance evaluation and an evidentiary audit.

11.2  Quality Assurance On-Site Evaluation.  Quality assurance evaluators
      inspect the Contractor's facilities to verify the adequacy and
      maintenance of instrumentation, the continuity of personnel meeting
      experience or education requirements, and the acceptable performance of
      analytical and QC procedures.

11.2.1   The Contractor shall expect that items to be monitored will include,
         but not be limited to, the following items:

         •   Size and appearance of the facility,

         •   Quantity, age, availability, scheduled maintenance and
             performance of instrumentation,

         •   Availability,  appropriateness, and utilization of the QAP and
             SOPs,

         •   Staff qualifications and experience, and personnel training
             programs,

         •   Reagents, standards, and sample storage facilities,

         •   Standard preparation logbooks and raw data,

         •   Bench sheets and analytical logbook maintenance and review, and

         •   Review of the Contractor's sample analysis/data package
             inspection/data management procedures.

11.2.2   Prior to an on-site evaluation, various documentation pertaining to
         performance of the specific Contractor is integrated in a profile
         package for discussion during the evaluation.   Items that may be
         included are previous on-site reports,  laboratory evaluation sample
         scores, Regional review of data, Regional QA materials,  GC/MS tape
         audit reports,  data audit reports, results of CCS, and date trend
         reports.

11.3  Evidentiary Audit.   Evidence auditors conduct an on-site laboratory
      evaluation to determine if Contractor policies and procedures are in
      place to satisfy evidence handling requirements as stated in Exhibit F.
      The evidence audit comprises a procedural audit,  an audit of written
      SOPs,  and an audit of analytical project file documentation.
                                     E-33                              OLM03.0

-------
Exhibit E — Section 11
On-Site Laboratory Evaluations
11.3.1   Procedural Audit.  The procedural audit consists of review and
         examination of actual standard operating procedures and accompanying
         documentation for the following Contractor operations:  sample
         receiving, sample storage, sample identification, sample security,
         sample tracking  (from receipt to completion of analysis) and
         analytical project file organization and assembly.

11.3.2   Written SOPs Audit.  The written SOPs audit consists of review and
         examination of the written SOPs to determine if they are accurate and
         complete for the following Contractor operations:  sample receiving,
         sample storage, sample identification, sample security, sample
         tracking (from receipt to completion of analysis) and analytical
         project file organization and assembly.

11.3.3   Analytical Project File Evidence Audit.  The analytical project file
         evidence audit consists of review and examination of the analytical
         project file documentation.   The auditors review the files to
         determine:

         •   The accuracy of the document inventory,

         •   The completeness of the  file,

         •   The adequacy and accuracy of the document numbering system,

         •   Traceability of sample activity,

         •   Identification of activity recorded on the documents,  and

         •   Error correction methods.

11.4  Discussion of the On-Site Team's Findings.  During the debriefing, the
      auditors present their findings and recommendations for corrective
      actions necessary to the Contractor personnel.

11.5  Corrective Action Reports for Follow-Through to Quality Assurance and
      Evidentiary Audit Reports.   Following an on-site laboratory evaluation,
      quality assurance and/or evidentiary audit reports which discuss
      deficiencies found during the on-site evaluation may be sent to the
      Contractor.   In a detailed letter,  the Contractor shall discuss the
      corrective actions implemented  to resolve the deficiencies discussed
      during the on-site evaluation and discussed in the report(s)  to the
      Technical Project Officer,  Administrative Project Officer, and EMSL/LV
      within 14 days of receipt of the report.

11.5.1   An alternate delivery schedule may be proposed by the Contractor, but
         it is the sole decision of the Agency, represented either by the
         Technical Project Officer or Administrative Project Officer, to
         approve or disapprove the alternate delivery schedule.  If an
         alternate delivery schedule  is proposed, the Contractor shall
         describe in a letter to the  Technical Project Officer, Administrative
         Project Officer,  and the Contracting Officer why he/she is unable to
         meet the delivery schedule listed in this section.  The Technical

                                     E-34                              OLM03.0

-------
                                                       Exhibit E — Section 11
                                                On-Site Laboratory Evaluations
         Project Officer/Administrative Project Officer will not grant an
         extension for greater than 14 days for the Contractor's response
         letter to the quality assurance and evidentiary audit report.  The
         Contractor shall proceed and not assume that an extension will be .
         granted until so notified by the Technical Project Officer and/or
         Administrative Project Officer.

11.5.2   If new SOPs are required to be written or SOPs are required to be
         amended because of the deficiencies and the subsequent corrective
         action implemented by the Contractor, the Contractor shall
         write/amend and submit the SOPs per the requirements listed in
         Section 4.

11.6  Corrective Actions.   If the Contractor fails to adhere to the
      requirements listed in Section 11, the Contractor may expect, but the
      Agency is not limited to, the following actions:   reduction in the
      number of samples sent under the contract,  suspension of sample shipment
      to the Contractor,  an on-site laboratory evaluation,  a GC/MS tape audit,
      a data package audit, a remedial laboratory evaluation sample, and/or
      contract sanctions,  such as a Cure Notice.
                                     E-35                               OLM03.0

-------
Exhibit E — Section 12
Quality Assurance and Data Trend Analysis
12.0  QUALITY ASSURANCE AND DATA TREND ANALYSIS

12.1  Data submitted by Contractors are subject to review from several
      aspects:  compliance with contract-required QC, usability, and full data
      package evaluation.  Problems resulting from any of these reviews may
      determine the need for a GC/MS tape audit, an on-site laboratory
      evaluation and/or a remedial laboratory evaluation sample.  In addition,
      QC prescribed in the methods provides information that is continually
      used by the Agency to assess sample data quality, Contractor data
      quality and Program data quality via data trend analysis.  Trend
      analysis is accomplished by entering data into a computerized database.
      Statistical reports that evaluate specific anomalies or disclose trends
      in many areas, including the following, are generated from this
      database:

      •   Surrogate spike recovery,

      •   Laboratory evaluation sample results,

      •   Blanks,

      •   GC/MS instrument performance checks (BFB and DFTPP),

      •   Initial and continuing calibration data, and

      •   Other QC and method parameters.

12.2  Program-wide statistical results are used to rank Contractors in order
      to observe the relative performance of each Contractor using a given
      protocol against its peers.   The reports are also used to identify
      trends within Contractors.  The results of many of these trend analyses
      are included in the overall  evaluation of a Contractor's performance,
      and are reviewed to determine if corrective action or an on-site
      laboratory evaluation may be required to ensure that the Contractor can
      meet the QA/QC requirements  of the contract.  Contractor performance
      over time is monitored using these trend analysis techniques to detect
      departures of Contractor output from required or desired levels of
      quality control, and to provide an early warning of Contractor QA/QC
      problems which may not be apparent from the results of an individual
      Case.

12.3  As a further benefit to the  Program, the database provides the
      information needed to establish performance-based criteria in updated
      analytical protocols, where  advisory criteria have been previously used.
      The vast empirical data set  produced by Contractors is carefully
      analyzed, with the results augmenting theoretical and research-based
      performance criteria.  The result is a continuously monitored set of
      quality control and performance criteria specifications of what is
      routinely achievable and expected of environmental chemistry Contractors
      engaged in mass production analysis of environmental samples.  This, in
      turn,  assists the Agency in  meeting its objectives of obtaining data of
      known and documented quality.
                                     E-36                              OLM03.0

-------
                                                       Exhibit E — Section 13
                                                               Data Management
 13.0  DATA MANAGEMENT

 13.1  Data management procedures are defined as procedures specifying the
      acquisition or entry, update, correction, deletion, storage and security
      of computer-readable data and files.  These procedures shall be in
      written  form and contain a clear definition for all databases and files
      used to  generate or resubmit deliverables.  Key areas of concern include
      system organization (including personnel and security), documentation
      operations, traceability and quality control.

 13.2  Data manually entered from hardcopy shall be subject to quality control
      and the  error rates estimated.  Systems shall prevent entry of incorrect
      or out-of-range data and alert data entry personnel of errors.  In
      addition, data entry error rates shall be estimated and recorded on a
      monthly  basis by reentering a statistical sample of the data entered and
      calculating discrepancy rates by data element.

 13.3  The record of changes in the form of corrections and updates to data
      originally generated,  submitted, and/or resubmitted shall be documented
      to allow traceability of updates.  Documentation shall include the
      following for each change.

      •   Justification or rationale for the change.

      •   Initials of the person making the change(s).  Data changes shall be
          implemented and reviewed by a person or group independent of the
          source generating the deliverable.

      •   Documentation of changes shall be retained according to the schedule
          of the original deliverable.

      •   Resubmitted diskettes or other deliverables shall be reinspected as
          a part of the Contractor's internal inspection process prior to
          resubmission.   The entire deliverable,  not just the changes,  shall
          be inspected.

      •   The Contractor's laboratory manager shall approve changes to
          originally submitted deliverables.

      •   Documentation of data changes may be requested by Contractor
          auditors.

13.4  Life cycle management  procedures shall  be applied to computer software
      systems developed by the Contractor to  be used to generate and edit
      contract deliverables.   Such systems shall be thoroughly tested and
      documented prior to utilization.

13.4.1   A software test and acceptance plan  including test requirements,  test
         results and acceptance criteria shall be developed,  followed,  and
         available in written form.

13.4.2   System changes  shall not be made directly to production systems
         generating deliverables.   Changes shall be made first to a
         development system  and tested prior  to implementation.

                                     E-37                              OLM03.0

-------
Exhibit E — Section 13
Data Management
13.4.3   Each version of the production system will be given an identification
         number, date of installation, date of last operation and archived.

13.4.4   System and operations documentation shall be developed and maintained
         for each system.  Documentation shall include a user's manual and an
         operations and maintenance manual.

13.4.5   This documentation shall be available for on-site review and/or upon
         written request by the TPO or APO.

13.5  Individual(s) responsible for the following functions shall be
      identified.

      •   System operation and maintenance, including documentation and
          training,

      •   Database integrity,  including data entry, data updating and quality
          control, and

      •   Data and system security,  backup and archiving.
                                     E-38                              OLM03.0

-------
                EXHIBIT F
   CHAIN-OF-CUSTODY, DOCUMENT CONTROL,
AND WRITTEN STANDARD OPERATING PROCEDURES
                   F-l                               OLM03.0

-------
             Exhibit F - Chain-of-Custody, Document Control, and
                     Written Standard Operating Procedures

                               Table  of  Contents
Section                                                                   Page

1.0   INTRODUCTION  	    3

2.0   STANDARD OPERATING PROCEDURES 	    4
      2.1   Sample Receiving  	    4
      2.2   Sample Identification 	    5
      2.3   Sample Security 	    5
      2.4   Sample Storage  	    6
      2.5   Sample Tracking and Document Control  	    6
      2.6   Computer-Resident Sample Data Control 	    7
      2.7   Complete Sample Delivery Group File (CSF) Organization and
            Assembly  	    7

3.0   WRITTEN STANDARD OPERATING PROCEDURES (SOPS)  	   10
      3.1   Sample Receiving  	   10
      3.2   Sample Identification 	   11
      3.3   Sample Security	   12
      3.4   Sample Storage	   12
      3.5   Sample Tracking and Document Control  	   12
      3.6   Computer-Resident Sample Data Control 	   13
      3.7   CSF Organization and Assembly	   14
                                     F-2                               OLM03.0

-------
                                                        Exhibit F — Section 1
                                                                  Introduction
1.0   INTRODUCTION

1.1   A sample is physical evidence collected from a facility or from the
      environment.  Controlling evidence is an essential part of the hazardous
      waste investigation effort.  To ensure that the Environmental Protection
      Agency's (EPA) sample data and records supporting sample-related
      activities are admissible and have weight as evidence in future
      litigation, Contractors are required to maintain EPA samples under
      chain-of-custody and to account for all samples and supporting records
      of sample handling, preparation, and analysis.  Contractors shall
      maintain sample identity, sample custody, and all sample-related records
      according to the requirements in this exhibit.

1.2   The purposes of the evidence requirements include:

      •   Ensuring traceability of samples while in the possession of the
          Contractor.

      •   Ensuring custody of samples while in the possession of the
          Contractor.

      •   Ensuring the integrity of sample identity while in the possession of
          the Contractor.

      •   Ensuring sample-related activities are recorded on documents or in
          other formats for EPA sample receipt, storage, preparation,
          analysis,  and disposal.

      •   Ensuring all laboratory records for each specified sample delivery
          group will be accounted for when the project is completed.

      •   Ensuring that all laboratory records directly related to EPA samples
          are assembled and delivered to EPA or, prior to delivery, are
          available upon EPA's request.
                                     F-3                               OLM03.0

-------
Exhibit F — Section 2
Standard Operating Procedures
2.0   STANDARD OPERATING PROCEDURES

      The Contractor shall implement the following standard operating
      procedures for sample receiving, sample identification, sample security,
      sample storage, sample tracking and document control, computer-resident
      sample data control, and complete sample delivery group file
      organization and assembly to ensure accountability of EPA sample
      chain-of-custody as well as control of all EPA sample-related records.

2.1   Sample Receiving

2.1.1    The Contractor shall designate a sample custodian responsible for
         receiving EPA samples.

2.1.2    The Contractor shall designate a representative to receive EPA
         samples in the event that the sample custodian is not available.

2.1.3    Upon receipt, the condition of shipping containers and sample
         containers shall be inspected and recorded on Form DC-1 by the sample
         custodian or his/her representative.

2.1.4    Upon receipt, the condition of the custody seals (intact/broken)
         shall be inspected and recorded on Form DC-1 by the sample custodian
         or his/her representative.

2.1.5    The sample custodian or his/her representative shall verify and
         record on Form DC-1 the presence or absence of the following
         documents accompanying the sample shipment:

         •   Custody seals,

         •   Chain-of-custody records,

         •   Traffic reports or packing lists,

         •   Airbills or airbill stickers,  and

         •   Sample tags.

2.1.6    The sample custodian or his/her representative shall verify and
         record on Form DC-1 the agreement or disagreement of information
         recorded on all documents received with samples and information
         recorded on sample containers.

2.1.7    The sample custodian or his/her representative shall record the
         following information on Form DC-1 as samples are received and
         inspected:

         •   Custody seal numbers when present,

         •   Airbill or airbill sticker numbers,

         •   Sample tags listed/not listed on chain-of-custody records,
                                      F-4                               OLM03.0

-------
                                                        Exhibit F — Section 2
                                                 Standard Operating Procedures
          •   Date of receipt,

          •   Time of receipt,

          •   EPA sample numbers,

          •   Sample tag numbers,

          •   Assigned laboratory numbers,

          •   Samples delivered by hand, and

          •   Problems and discrepancies.

2.1.8    The sample custodian or his/her representative shall sign, date, and
         record the time on all accompanying forms, when applicable, at the
         time of sample receipt (for example, chain-of-custody records,
         traffic reports or packing lists, and airbills).  Note: Initials are
         not acceptable.

2.1.9    The Contractor shall contact the Sample Management Office  (SMO) to
         resolve problems and discrepancies including but not limited to,
         absent documents, conflicting information, absent or broken custody
         seals, and unsatisfactory sample condition (for example, leaking
         sample container).

2.1.10   The Contractor shall record resolution c: problems and discrepancies
         by SMO.

2.2   Sample Identification

2.2.1    The Contractor shall maintain the identity of EPA samples and
         prepared samples (including extracted samples, digested samples, and
         distilled samples) throughout the laboratory.

2.2.2    Each sample and sample preparation container shall be labeled with
         the SMO number or a unique laboratory sample identification number.

2.3   Sample Security

2.3.1    The Contractor shall demonstrate that EPA sample custody is
         maintained from receiving through retention or disposal.  A sample is
         in custody if:

         •   It is in your possession; or

         •   It is in your view after being in your possession; or

         •   It is locked in a secure area after being in your possession; or

         •   It is in a designated secure area.  (Secure areas shall be
             accessible only to authorized personnel.)

2.3.2    The Contractor shall demonstrate security of designated secure areas.

                                      F-5                               OLM03.0

-------
 Exhibit F — Section 2
 Standard Operating Procedures
 2.4   Sample Storage

       The Contractor shall designate storage areas for EPA samples and
       prepared samples.

 2.5   Sample Tracking and Document Control

 2.5.1    The Contractor  shall record all activities performed on EPA samples.

 2.5.2    Titles which identify the activities recorded shall  be printed on
          each page of all laboratory documents.   (Activities  include,  but are
          not limited to,  sample receipt,  sample  storage,  sample preparation,
          and sample analysis.)   When a  document  is a record of analysis,  the
          instrument type  and parameter  group (for  example,  GC/MS-VOA)  shall be
          included in the  title.

 2.5.3    When columns are used  to  organize information recorded on laboratory
          documents,  the information  recorded in  the columns shall  be
          identified in a  column heading.

 2.5.4    Reviewers'  signatures  shall  be  identified on  laboratory documents
          when reviews are conducted.  Note:  Individuals recording  review
          comments on computer-generated  raw  data are not  required  to be
          identified unless the  written  comments  address data  validity.

 2.5.5     The laboratory name shall be identified on preprinted laboratory
          documents.

 2.5.6     Each laboratory  document entry  shall be dated with the month/day/year
          (for example, 01/01/90) and  signed  (or  initialed)  by the
          individual(s) responsible for performing  the  recorded activity at the
          time the activity is recorded.

 2.5.7     Notations  on  laboratory documents shall be recorded  in ink.

 2.5.8     Corrections  to laboratory documents and raw data shall  be made by
          drawing  single lines through the errors and entering the  correct
          information.  Information shall not be obliterated or rendered
          unreadable.  Corrections and additions to  information shall be signed
          (or  initialed) and  dated.

 2.5.9     Unused portions  of  laboratory documents shall be lined-out.

 2.5.10    Pages in bound and  unbound logbooks shall be sequentially numbered.

2.5.11    Instrument-specific run logs shall be maintained to  enable  the
          reconstruction of run sequences.

2.5.12    Logbook  entries  shall be in chronological order.

2.5.13    Logbook entries  shall include only one Sample Delivery Group  (SDG)
         per page, except  in the events where the SDGs "share" QC samples  (for
          example, instrument run logs and extraction logs).
                                      F-6                               OLM03.0

-------
                                                        Exhibit P — Section 2
                                                 Standard Operating Procedures
2.5.14    Information  inserted  into laboratory documents shall be affixed
          permanently  in place.  The  individual responsible for inserting
          information  shall sign and  date across the  insert and logbook page at
          the time  information  is inserted.

2.5.1b    The Contractor shall  document disposal or retention of EPA samples,
          remaining portions of samples, and prepared samples.

2.6   Computer-Resident Sample Data  Control

2.6.1     Contractor personnel  responsible for original data entry shall be
          identified at the time of data input.

2.6.2     The Contractor shall make changes to electronic data in a manner
          which ensures that the original data entry  is preserved, the editor
          is identified, and the revision date is recorded.

2.6.3     The Contractor shall routinely verify the accuracy of manually
          entered data, electronically entered data, and data acquired from
          instruments.

2.6.4     The Contractor shall routinely verify documents produced by the
          electronic data collection  system to ensure accuracy of the
          information reported.

2.6.5     The Contractor shall ensure that the electronic data collection
          system  s secure.

2.6.5.1     The electronic data collection system shall be maintained in a
            secure location.

2.6.5.2     Access to the electronic data collection system functions shall be
            limited to authorized personnel through utilization of software
            security techniques (for example, log-ons or restricted
            passwords).

2.6.5.3     Electronic data collection systems shall be protected from the
            introduction of external programs or software (for example,
            viruses).

2.6.6    The Contractor shall designate archive storage areas for electronic
         data and the software required to access the data.

2.6.7    The Contractor shall designate an individual responsible for
         maintaining archives of electronic data including the software.

2.6.8    The Contractor shall maintain the archives of electronic data and
         necessary software in a secure location.   (Secure areas shall be
         accessible only to authorized personnel.)

2.7    Complete Sample Delivery Group File (CSF)  Organization and Assembly

2.7.1    The Contractor shall designate a document control officer responsible
         for the organization and assembly of the CSF.

                                      F-7                               OLM03.0

-------
Exhibit F — Section 2
Standard Operating Procedures


2.7.2    The Contractor shall designate a representative responsible for the
         organization and assembly of the CSF in the event that the document
         control officer is not available.

2.7.3    The Contractor shall maintain documents relating to the CSF in a
         secure location.

2.7.4    All original laboratory forms and copies of SDG-related logbook pages
         shall be included in the CSF.

2.7.5    Copies of laboratory documents in the CSF shall be photocopied in a
         manner to provide complete and legible replicates.

2.7.6    Documents relevant to each SDG including, but not limited to, the
         following shall be included in the CSF:

         •  logbook pages,             •  records of failed or attempted analysis,
         •  benchsheets,               •  custody records,
         •  mass spectra,              •  sample tracking records,
         •  chromatograms,             •  raw data summaries,
         •  screening records,         •  computer printouts,
         •  preparation records,       •  correspondence,
         •  re-preparation records,    •  FAX originals,
         •  analytical records,        •  library search results, and
         •  re-analysis records,       •  other.

2.7.7    The document control officer or his/her representative shall ensure
         that sample tags are encased in clear plastic bags before placing
         them in the CSF.

2.7.8    CSF documents shall be organized and assembled on an SDG-specific
         basis.

2.7.9    Original documents which include information relating to more than
         one SDG  (for example, chain-of-custody records, traffic reports,
         calibration logs) shall be filed in the CSF of the lowest SDG number,
         and copies of these originals shall be placed in the other CSF(s).
         The document control officer or his/her representative shall record
         the following statement on the copies in dark ink:
                                     COPY
                ORIGINAL  DOCUMENTS ARE  INCLUDED  IN CSF
                                                           Signature
                                                            Date
                                     F-8                               OLM03.0

-------
                                                        Exhibit F — Section 2
                                                 Standard Operating Procedures
2.7.10   All CSFs shall be submitted with a completed Form DC-2.  All
         resubmitted CSFs shall be submitted with a new or revised Form DC-2.

2.7.11   Each item in the CSF and resubmitted CSFs shall be inventoried and
         assembled in the order specified on Form DC-2.  Each page of the CSF
         shall be stamped with a sequential number.  Page number ranges shall
         be recorded in the columns provided on Form DC-2.  Intentional gaps
         in the page numbering sequence shall be recorded in the "Comments"
         section on Form DC-2.  When inserting new or inadvertently omitted
         documents, the Contractor shall identify them with unique accountable
         numbers.  The unique accountable numbers and the locations of the
         documents shall be recorded in the "Other Records" section on Form
         DC-2.

2.7.12   Before shipping each CSF, the document control officer or his/her
         representative shall verify the agreement of information recorded on
         all documentation and ensure that the information is consistent and
         the CSF is complete.

2.7.13   The document control officer or his/her representative shall document
         the shipment of deliverable packages including what was sent, to
         whom, the date, and the carrier used.

2.7.14   Shipments of deliverable packages,  including resubmittals, shall be
         sealed with custody seals by the document control officer or his/her
         representative in a manner such that opening the packages would break
         the seals.

2.7.15   Custody seals shall be signed and dated by the document control
         officer or his/her representative when sealing deliverable packages.
                                      F-9                               OLM03.0

-------
Exhibit F — Section 3
Written Standard Operating Procedures
3.0   WRITTEN STANDARD OPERATING PROCEDURES (SOPS)

      The Contractor shall develop and implement the following written SOPs
      for sample receiving, sample identification, sample security, sample
      storage, sample tracking and document control, computer-resident sample
      data control, and CSF file organization and assembly to ensure
      accountability for EPA sample chain-of-custody and control of all EPA
      sample-related records.

3.1   Sample Receiving

3.1.1    The Contractor shall have written SOPs for sample receiving which
         accurately reflect the procedures used by the laboratory.

3.1.2    The written SOPs for sample receiving shall ensure that the
         procedures listed below are in use at the laboratory.

3.1.2.1     The condition of shipping containers and sample containers are
            inspected and recorded on Form DC-1 upon receipt by the sample
            custodian or his/her representative.

3.1.2.2     The condition of custody seals are inspected and recorded on Form
            DC-1 upon receipt by the sample custodian or uis/her
            representat ive.

3.1.2.3     The presence or absence of the following documents accompanying
            the sample shipment is verified and recorded on Form DC-1 by the
            sample custodian or his/her representative:

            •   Custody seals,

            •   Chain-of-custody records,

            •   Traffic reports or packing lists,

            •   Airbills or airbill stickers,  and

            •   Sample tags.

3.1.2.4     The agreement or disagreement of information recorded on shipping
            documents with information recorded on sample containers is
            verified and recorded on Form DC-1 by the sample custodian or
            his/her representative.

3.1.2.5     The following information is recorded on Form DC-1 by the sample
            custodian or his/her representative as samples are received and
            inspected:

            •   Custody seal numbers when present,

            •   Airbill or airbill sticker numbers,

            •   Sample tag numbers listed/not listed on chain-of-custody
                records,

                                     F-10                              OLM03.0

-------
                                                        Exhibit  F  —  Section  3
                                         Written Standard Operating Procedures
             •   Date  of  receipt,

             •   Time  of  receipt,

             •   EPA sample numbers,

             •   Sample tag numbers,

             •   Assigned laboratory numbers,

             •   Samples  delivered by hand, and

             •   Problems and discrepancies.

3.1.2.6      All accompanying forms are signed, dated, and the time is
             recorded, when applicable, at the time of sample receipt  (for
             example,  chain-of-custody records, traffic reports or packing
             lists, and airbills) by the sample custodian or his/her
             representative.

3.1.2.7      SMO is contacted to resolve problems and discrepancies including,
             but not limited to, absent documents, conflicting information,
             absent or broken custody seals, and unsatisfactory sample
             condition (for example, leaking sample container).

3.1.2.8      The resolution of problems and discrepancies by SMO is recorded.

3.2   Sample Identification

3.2.1    The Contractor  shall have written SOPs for sample identification
         which accurately reflect the procedures used by the laboratory.

3.2.2    The written SOPs for sample identification shall ensure that the
         procedures listed below are in use at the laboratory.

3.2.2.1     The identity of EPA sarples and prepared samples is maintained
            throughout the laboratory:

             •   When the Contractor assigns unique laboratory sample
                identification numbers, the written SOPs shall include a
                description of the procedure used to assign these numbers,

             •   When the Contractor uses prefixes or suffixes in addition to
                laboratory sample identification numbers, the written SOPs
                shall include their definitions,  and

             •   When the Contractor uses methods to uniquely identify
                fractions/parameter groups and matrix type,  the written SOPs
                shall include a description of these methods.

3.2.2.2     Each sample and sample preparation container is labeled with the
            SMO number or a unique laboratory sample identification number.
                                     F-ll                              OLM03.0

-------
Exhibit F — Section 3
Written Standard Operating Procedures
3.3   Sample Security

3.3.1    The Contractor shall have written SOPs for sample security which
         accurately reflect the procedures used by the laboratory.

3.3.2    The written SOPs for sample security shall include the items listed
         below.

3.3.2.1     Procedures which ensure the following:

            •   Sample custody is maintained, and

            •   The security of designated secure areas is maintained.

3.3.2.2     A list of authorized personnel who have access to locked storage
            areas.

3.4   Sample Storage

3.4.1    The Contractor shall have written SOPs for sample storage which
         accurately reflect the procedures used by the laboratory.

3.4.2    The written SOPs for sample storage shall describe locations,
         contents, and identities of all storage areas for EPA samples and
         prepared samples in the laboratory.

3.5   Sample Tr ;king and Document Control

3.5.1    The Contractor shall have written SOPs for sample tracking and
         document control which accurately reflect the procedures used by the
         laboratory.

3.5.2    The written SOPs for sample tracking and document control shall
         include the items listed below.

3.5.2.1     Examples of all laboratory documents used during sample receiving,
            sample storage,  sample transfer, sample analyses, CSF organization
            and assembly, and sample retention or disposal.

3.5.2.2     Procedures which ensure the following:

            •   All activities performed on EPA samples are recorded;

            •   Titles which identify the activities recorded are printed on
                each page of all laboratory documents;

            •   Information recorded in columns is identified with column
                headings;

            •   Reviewers' signatures are identified on laboratory documents;

            •   The laboratory name is included on preprinted laboratory
                documents;
                                     F-12                              OLM03.0

-------
                                                        Exhibit F  — Section 3
                                         Written  Standard Operating Procedures
             •    Laboratory document  entries  are  signed  and  dated with the
                 month/day/year  (for  example, 01/01/90);

             •    Entries  on all  laboratory documents are recorded in  ink;

             •    Corrections  and additions to laboratory documents are made by
                 drawing  single  lines through the errors, entering the correct
                 information, and  initialing  and  dating the  new  information;

             •    Unused portions of laboratory documents are lined-out;

             •    Pages in bound  and unbound logbooks are sequentially numbered;

             •    Instrument-specific  run logs are maintained to  enable the
                 reconstruction  of run sequences;

             •    Logbook  entries are  recorded in  chronological order;

             •    Entries  are  recorded for only one SDG on a  page, except in the
                 events where SDGs "share" quality control (QC)  samples (for
                 example, instrument  run logs and extraction logs);

             •    Information  inserted in laboratory documents is affixed
                 permanently, signed, and dated across the insert; and

             •    The retention or disposal of EPA samples, remaining  portions
                 of samples,  and prepared samples is documented.

3.6   Computer-Resident  Sample  Data  Control

3.6.1    The Contractor  shall have written SOPs  for computer-resident sample
         data control which  accurately reflect the procedures used by the
         laboratory.

3.6.2    The written SOPs for computer-resident  sample data control  shall
         include the items listed below.

3.6.2.1     Procedures which ensure  the following:

            •   Contractor personnel responsible for original data entry are
                 identified;

            •   Changes to electronic data are made such that the original
                data entry is preserved, the editor is identified,  and the
                revision date is recorded;

            •   The accuracy of manually entered data, electronically entered
                data, and data  acquired from instruments is verified;

            •   Report documents produced by the electronic data collection
                system are routinely verified to ensure the accuracy of the
                 information reported;

            •   Electronic data collection system security  is maintained; and

                                     F-13                              OLM03.0

-------
Exhibit F — Section 3
Written Standard Operating Procedures
            •   Archives of electronic data and accompanying software are
                maintained in a secure location.

3.6.2.2     Descriptions of archive storage areas for the electronic data and
            the software required to access data archives.

3.6.2.3     A list of authorized personnel who have access to electronic data
            collection system functions and to archived data.

3.7   CSF Organization and Assembly

3.7.1    The Contractor shall have written SOPs for CSF organization and
         assembly which accurately reflect the procedures used by the
         laboratory.

3.7.2    The written SOPs for CSF organization and assembly shall ensure that
         the procedures listed below are in use at the laboratory.

         •   Documents relating to the CSF are maintained in a secure
             location.

         •   All original laboratory forms and copies of SDG-related logbook
             pages are included in the CSF.

         •   Laboratory documents are photocopied in a manner to provide
             complete and legible replicates.

         •   All documents relevant to each SDG are included in the CSF.

         •   Sample tags are encased in clear  plastic bags by the document
             control officer or his/her representative before placing them in
             the CSF.

         •   The CSF is organized and assembled on an SDG-specific basis.

         •   Copies are referenced to originals in the event that an original
             document contains information relating to more than one SDG.

         •   Each CSF is submitted with a completed Form DC-2, and resubmitted
             CSFs are submitted with a new or  revised Form DC-2.

         •   Each page of the CSF is stamped with a sequential number and the
             page number ranges are recorded in the columns provided on Form
             DC-2.

         •   Consistency and completeness of the CSF is verified by the
             document control officer or his/her representative.

         •   Shipments of deliverable packages are documented by the document
             control officer or his/her representative.

         •   Deliverable packages are shipped  by the document control officer
             or his/her representative using custody seals in a manner such
             that opening the packages would break the seals.

                                     F-14                              OLM03.0

-------
                                               Exhibit F — Section 3
                                Written Standard Operating Procedures
•   Custody seals are signed and dated by the document control
    officer or his/her representative before placing them on
    deliverable packages.
                            F-15                               OLM03.0

-------
    EXHIBIT G
GLOSSARY OF TERMS
       G-l
OLM03.0

-------
Exhibit G — Glossary of Terms
ALIQUOT - a measured portion of a sample, or solution, taken for sample
preparation and/or analysis.

ANALYSIS DATE/TIME - the date and military time of the injection of the
sample, standard, or blank into the GC/MS or GC system.

BAR GRAPH SPECTRUM - a plot of the mass-to-charge ratio (m/e) versus relative
intensity of the ion current.

BLANK - an analytical sample designed to assess specific sources of laboratory
contamination.  See individual types of Blanks:  Method Blank; Instrument
Blank, Storage Blank, and Sulfur Blank.

BREAKDOWN - a measure of the decomposition of certain analytes (DDT and
Endrin) into by-products.

4-BROMOFLUOROBENZENE (BFB) - the compound chosen to establish mass spectral
instrument performance for volatile (VOA) analyses.  It is also used in the
VOA fraction as a system monitoring compound (SMC).

CALIBRATION FACTOR (CF) - a measure of the gas chromatographic response of a
target analyte to the mass injected.  The calibration factor is analogous to
the Relative Response Factor (RRF) used in the Volatile and Semivolatile
fractions.

CASE - a finite, usually predetermined number of samples collected over a
given time period from a particular site.  Case numbers are assigned by the
Sample Management Office.  A Case consists of one or more Sample Delivery
Groups.

CHARACTERIZATION - a determination of the approximate concentration range of
compounds of interest used to choose the appropriate analytical protocol.

CONCENTRATION LEVEL (low or medium)  - characterization of soil samples or
sample fractions as low concentration or medium concentration is made on the
basis of the laboratory's preliminary screen, not on the basis of information
entered on the Traffic Report by the sampler.

CONTAMINATION - a component of a sample or an extract that is not
representative of the environmental source of the sample.   Contamination may
stem from other samples, sampling equipment, while in transit, from laboratory
reagents,  laboratory environment,  or analytical instruments.

CONTINUING CALIBRATION - analytical standard run every 12 hours to verify the
initial calibration of the system.

CONTINUOUS LIQUID-LIQUID EXTRACTION - used herein synonymously with the terms
continuous extraction,  continuous liquid extraction, and liquid extraction.
This extraction technique involves boiling the extraction solvent in a flask
and condensing the solvent above the aqueous sample.  The condensed solvent
drips through the sample, extracting the compounds of interest from the
aqueous phase.

DATE - MM/DD/YY - where MM = 01 for January, 02 for February, ... 12 for
December;  DD = 01 to 31; YY = 94,  95,  96, 97, etc.

                                      G-2                               OLM03.0

-------
                                                Exhibit G — Glossary of Terms
DAY - unless otherwise  specified, day  shall mean calendar day.

DECAFLUOROTRIPHENYLPHOSPHINE  (DFTPP) - compound chosen to establish mass
spectral  instrument performance  for  semivolatile analysis.

EXTRACTABLE - a compound that  can be partitioned into an organic  solvent  from
the sample matrix and is amenable to ga^ chromatography.  Extractables  include
semivolatile (BNA) and  pesticide/Aroclor compounds.

EXTRACTED ION CURRENT PROFILE  (EICP) - a plot of ion abundance versus time  (or
scan number) for ion(s) of specified mass(es).

GAS CHROMATOGRAPH (GC)  - the instrument used to separate analytes on a
stationary phase within a chromatographic column.  The analytes are volatized
directly  from the sample (VOA  water  and low-soil), volatized from the sample
extract (VOA medium soil), or  injected as extracts (SVOA and PEST).  In VOA
and SVOA  analysis, the  compounds are detected by a Mass Spectrometer (MS).  In
PEST analysis,  the compounds are detected by an Electron Capture  (EC)
detector.  In the screening procedure  (all fractions), the Flame  lonization
Detector  (FID)  is used  as the  detector.

GEL PERMEATION CHROMATOGRAPHY  (GPC)  -  a size-exclusion chromatographic
technique that is used  as a cleanup  procedure for removing large organic
molecules, particularly naturally occurring macro-molecules such as lipids,
polymers, viruses, etc.

IN-HOUSE - at the Contractor's facility.

INITIAL CALIBRATION - analysis of analytical standards for a series of
different specified concentrations;  used to define the linearity and dynamic
range of the response of the mass spectrometer or electron capture detector to
the target compounds.

INTEGRATION SCAN RANGE  - the scan number of the scan at the beginning of the
area of integration to  the scan number at the end of the area of integration.
Performed in accordance with Exhibit D VOA, Sections 11.2.1.9 and 11.2.1.10
and Exhibit D SVOA,  Sections 11.2.1.2  and 11.2.1.3.

INTEGRATION TIME RANGE  - the retention time at the beginning of the area of
integration to the retention time at the end of the area of integration.

INTERNAL STANDARDS - compounds added to every standard,  blank, matrix spike,
matrix spike duplicate, sample (for  volatiles), and sample extract (for
semivolatiles)  at a known concentration, prior to analysis.  Internal
standards are used as the basis for  quantitation of the target compounds.

INSTRUMENT BLANK - a blank designed  to determine the level of contamination
associated with the analytical instruments.

INSUFFICIENT QUANTITY - when there is  not enough volume (water sample) or
weight (soil/sediment)  to perform any  of the required operations:  sample
analysis or extraction, percent moisture,  MS/MSD, etc. Exhibit D provides
guidance for addressing this situation.

LABORATORY - synonymous with Contractor as used herein.

m/z - Mass to charge ratio, synonymous with "m/e".


                                     G-3                               OLM03.1

-------
 Exhibit G  — Glossary of Terms
MATRIX - the predominant material of which the sample to be analyzed  is
composed. For the purpose of this SOW, a sample matrix is either water or
soil/sediment. Matrix is not synonymous with phase  (liquid or solid).

MATRIX EFFECT -  in general, the effect of a particular matrix  (water or
soil/sediment) on the constituents with which it contacts.  This is
particularly pronounced for clay particles which may adsorb chemicals and
catalyze reactions.  Matrix effects may prevent extraction of target  analytes,
and may affect surrogate recoveries.  In addition, nontarget analytes may be
extracted from the matrix causing interferences.

MATRIX SPIKE - aliquot of a matrix (water or soil) fortified (spiked) with
known quantities of specific compounds and subjected to the entire analytical
procedure in order to indicate the appropriateness of the method for  the
matrix by measuring recovery.

MATRIX SPIKE DUPLICATE - a second aliquot of the same matrix as the matrix
spike (above) that is spiked in order to determine the precision of the
method.

METHOD BLANK - an analytical control consisting of all reagents, internal
standards and surrogate standards (or SMCs for VOA), that is carried
throughout the entire analytical procedure.  The method blank is used to
define the level of laboratory, background and reagent contamination.

NARRATIVE (SDG Narrative)  - portion of the data package which includes
laboratory,  contract, Case and sample number identification, and descriptive
documentation of any problems encountered in processing the samples,  along
with corrective action taken and problem resolution.  Complete SDG Narrative
specifications are included in Exhibit B.

PERCENT DIFFERENCE (%D)  - As used in this SOW and elsewhere to compare two
values,  the percent difference indicates both the direction and the magnitude
of the comparison,  i.e.,  the percent difference may be either negative,
positive,  or zero.  (In contrast, see relative percent difference.)

PERCENT MOISTURE - an approximation of the amount of water in a ^oil/sediment
sample made by drying an aliquot of the sample at 105 °C.   The percent
moisture determined in this manner also includes contributions from all
compounds that may volatilize at or below 105 °C,  including water.   Percent
moisture may be determined from decanted samples and from samples that are not
decanted.

PERFORMANCE EVALUATION MIXTURE - a calibration solution of specific analytes
used to evaluate both recovery and percent breakdown as measures of
performance.

PRIMARY QUANTITATION ION - a contract specified ion used to quantitate a
target analyte.

PROTOCOL - describes the exact procedures to be followed with respect to
sample receipt and handling, analytical methods, data reporting and
deliverables, and document control.   Used synonymously with Statement of Work
(SOW).
                                      G-4                               OLM03.0

-------
                                                Exhibit G  — Glossary  of Terms
PURGE AND TRAP  (DEVICE)  -  analytical technique  (device) used to  isolate
volatile  (purgeable) organics by stripping the  compounds  from water or soil by
a stream of inert gas, trapping the compounds on an  adsorbent such as a porous
polymer trap, and thermally desorbing the trapped compounds onto the gas
chromatographic column.

PURGEABLES - volatile compounds.

REAGENT WATER - water in which an interferant is not observed at or above the
minimum quantitation limit of the parameters of interest.

RECONSTRUCTED ION CHROMATOGRAM (RIC) - a mass spectral graphical
representation of the separation achieved by a  gas chromatograph; a plot of
total ion current versus retention time.

RELATIVE PERCENT DIFFERENCE (RPD) - As used in  this  SOW and elsewhere to
compare two values, the  relative percent difference  is based on  the mean of
the two values, and is reported as an absolute  value, i.e., always expressed
as a positive number or  zero.  (In contrast, see percent  difference.)

RELATIVE RESPONSE FACTOR (RRF) - a measure of the relative mass  spectral
response of an analyte compared to its internal standard.  Relative Response
Factors are determined by analysis of standards and  =>.re used in  the
calculation of concentrations of analytes in samples.  RRr" is determined by
the following equation:


                                       A    C
                               RRF  =  _2* X -12
                                      "is    x

      Where,

      A  = area of the characteristic ion measured
      C  = concentration, or amount (mass)
      is = internal standard
      x  = analyte of interest

RELATIVE RETENTION TIME  (RRT)  - the ratio of the retention time  of a compound
to that of a standard (such as an internal standard).


                                         RTr
                                  RRT =
      Where,

      RTC = Retention time for the semivolatile target or surrogate compound
            in continuing calibration.
      RTjs=  Retention time for the internal standard in calibration standard
            or in a sample.
                                      G-5                               OLM03.0

-------
Exhibit G — Glossary of Terms
RESOLUTION - also termed separation or percent resolution, the separation
between peaks on a chromatogram, calculated by dividing the depth of the
valley between the peaks by the peak height of the smaller peak being
resolved, multiplied by 100.
                       ATM (or Hdgtlt)
For pesticide analysis the X-axis shall be displayed such that a data reviewer
can calculate the % Resolution.

RESOLUTION CHECK MIXTURE - a solution of specific analytes used to determine
resolution of adjacent peaks; used to assess instrumental performance.

RESPONSE - or Instrumental Response: a measurement of the output of the GC
detector (MS, EC, or FID) in which the intensity of the signal is
proportionate to the amount  (or concentration) detected.  Measured by peak
area or peak height.

RETENTION TIME (RT) - the time a target analyte is retained on a GC column
before elution.  The identification of a target analyte is dependent on a
target compound's retention time falling within the specified retention time
window established for that compound.  Retention time is dependent on the
nature of the column's stationary phase, column diameter, temperature, flow
rate,  and other parameters.

SAMPLE - a portion of material to be analyzed that is contained in single or
multiple containers and identified by a unique sample number.

SAMPLE DELIVERY GROUP (SDG) - a unit within a single Case that is used to
identify a group of samples for delivery.  An SDG is a group of 20 or fewer
field samples within a Case, received over a period of up to 14 calendar days
(7 calendar days for 14-day data turnaround contracts).  Data from all samples
in an SDG are due concurrently.  A Sample Delivery Group is defined by one of
the following, whichever occurs first:

•     All samples within a Case; or

•     Every set of 20 field samples within a Case; or
                                      G-6
OLM03.0

-------
                                                Exhibit G — Glossary of Terms
•     All samples received within a 14-day calendar period  (7-day calendar
      period for 14-day data turnaround contracts).

Samples may be assigned to Sample Delivery Groups by matrix  (i.e., all soil
samples in one SDG, all water samples in another), at the discretion of the
laboratory.

SAMPLE NUMBER (EPA Sample Number) - a unique identification number designated
by EPA to each sample.  The EPA sample number appears on the sample Traffic
Report which documents information on that sample.

SECONDARY QUANTITATION ION - contract specified ion(s) to be used in
quantitation of target analytes when interferences prevent the use of the
primary guantitation ion.

SEMIVOLATILE COMPOUNDS - compounds amenable to analysis by extraction of the
sample with an organic solvent.  Used synonymously with Base/Neutral/Acid
(BNA) compounds.

SOIL - used herein synonymously with soil/sediment and sediment.

SONIC CELL DISRUPTOR (SONICATOR) - a device that uses the energy from
controlled ultrasound applications to mix, disperse, and dissolve organic
materials from a given matrix.

STANDARD ANALYSIS - an analytical determination made with known quantities of
target compounds; used to determine response factors.

STORAGE BLANK - reagent water (two 40.0 mL aliquots) stored with samples in an
SDG.  It is analyzed after all samples in that SDG have been analyzed; and is
used to determine the level of contamination acquired during storage.

SULFUR BLANK - a modified method blank that is prepared only when some of the
samples in a batch are subjected to sulfur cleanup.  It is used to determine
the level of contamination associated with the sulfur cleanup procedure.  When
all of the samples are subjected to  sulfur cleanup, then the method blank
serves this purpose.  When none of the samples are subjected to sulfur
cleanup,  no sulfur blank is required.

SURROGATES (Surrogate Standard) - for semivolatiles and pesticides/Aroclors,
compounds added to every blank, sample,  matrix spike, matrix spike duplicate,
and standard; used to evaluate analytical efficiency by measuring recovery.
Surrogates are brominated, fluorinated,  or isotopically labelled compounds not
expected to be detected in environmental media.

SYSTEM MONITORING COMPOUNDS - compounds added to every blank, sample, matrix
spike, matrix spike duplicate, and standard for volatile analysis, and used to
evaluate the performance of the entire purge and trap-gas chromatograph-mass
spectrometer system.  These compounds are brominated or deuterated compounds
not expected to be detected in environmental media.

TARGET COMPOUND LIST (TCL) - a list of compounds designated by the Statement
of Work (Exhibit C)  for analysis.
                                      G-7                               OLM03.0

-------
Exhibit G — Glossary of Terms
TENTATIVELY IDENTIFIED COMPOUNDS  (TIC) - compounds detected in  samples that
are not target compounds, internal standards, system monitoring compounds, or
surrogates.  Up to 30 peaks not including those tentatively identified as
alkanes,  (those greater than 10% of the peak areas or heights of nearest
internal  standard) are subjected to mass spectral library searches for
tentative identification.

TIME - when required to record time on any deliverable item, time shall be
expressed as Military Time, i.e., a 24-hour clock.

TRAFFIC REPORT (TR) - an EPA sample identification form filled out by the
sampler, which accompanies the sample during shipment to the laboratory and
which documents sample condition and receipt by the laboratory.

TWELVE-HOUR TIME PERIOD - The twelve (12) hour time period for GC/MS system
instrument performance check, standards calibration (initial or continuing
calibration), and method blank analysis begins at the moment of injection of
the DFTPP or BFB analysis that the laboratory submits as documentation of
instrument performance.  The time period ends after 12 hours have elapsed
according to the system clock.  For pesticide/Aroclor analyses performed by
GC/EC, the twelve hour time period in the analytical sequence begins at the
moment of injection of the instrument blank that precedes sample analyses, and
ends after twelve hours have elapsed according to the system clock.

VALIDATED TIME OF SAMPLE RECEIPT (VTSR) - the date on which a sample is
received at the Contractor's facility,  as recorded on the shipper's delivery
receipt and Sample Traffic Report.

VOLATILE COMPOUNDS - compounds amenable to analysis by the purge and trap
technique.  Used synonymously with purgeable compounds.

WIDE BORE CAPILLARY COLUMN - a gas chromatographic column with an internal
diameter  (ID) that is greater than or equal to 0.53 mm.   Columns with lesser
diameters are classified as narrow bore capillary columns.
                                      G-8                               OLM03.1

-------
           EXHIBIT  H
AGENCY STANDARD IMPLEMENTATION
             H-l                               OLM03.0

-------
                  Exhibit H - Agency Standard  Implementation

                               Table of Contents


Section                                                                    Page


1.0   FORMAT CHARACTERISTICS  	     3

2.0   RECORD TYPES   	     4

3.0   PRODUCTION RUNS	     5

4.0   RECORD SEQUENCE  	     7

5.0   FILE/RECORD INTEGRITY 	     8

6.0   DATES AND TIMES	     8

7.0   MULTIPLE VOLUME DATA	     8

8.0   DELIVERABLE 	     9

9.0   RECORD LISTING	    11
      9.1   Production Run Header Record  (Type 10)   	    11
      9.2   Chromatography Record (Type 11)	    12
      9.3   Sample Header Data Record  (Type 20)	    13
      9.4   Sample Header Data Record  (Type 21)	    15
      9.5   Sample Condition Record (Type 22)  	    16
      9.6   Associated Injection and Counter Record  (Type 23)  	    18
      9.7   Sample Cleanup Record (Type 27) 	    21
      9.8   Results Data Record (Type  30)	    23
      9.9   Auxiliary Data Record (Type 32)	    26
      9.10  Name Record (Type 33)	    27
      9.11  Instrumental Data Readout  Record (Type 36)   	    28
      9.12  Comment Record (Type 90)   	    31

10.0  DEFINITIONS OF VARIOUS CODES USED IN AGENCY STANDARD RECORDS   ...    32
      10.1  Quality Control and Related Codes  (QCC)  in Type 20 Records   .    32
      10.2  Codes For Sample Medium (Matrix, Sources) 	    34
      10.3  List of Sample and Result  Qualifiers	    34

APPENDIX A — FORMAT OF RECORDS FOR SPECIFIC USES	36
      Table of Contents	37
                                      H-2                               OLM03.0

-------
                                                        Exhibit H — Section 1
                                                        Format Characteristics
1.0   FORMAT CHARACTERISTICS

1.1   This constitutes an implementation of the EPA Agency Standard for
      Electronic Data Transmission based upon analytical results and ancillary
      information required by the contract.  All data generated by a single
      analysis are grouped together, and the groups are aggregated to produce
      files that report data from an SDG.  Because this implementation is only
      a subset of the Agency Standard, some fields have been replaced by
      delimiters as place holders for non-CLP data elements.

1.2   This implementation includes detailed specifications for the required
      format of each record.  The position in the record where each field is
      to be contained relevant to other fields is specified, as well as the
      maximum length of the field.  Each field's required contents are
      specified as literal (contained in quotes), which must appear exactly as
      shown (without quotes), or as a variable for which format and/or
      descriptions are listed in the format/contents column.  Options and
      examples are listed for most fields.  For fields where more than three
      options are available, a list and description of options are supplied on
      a separate page following the record descriptions.  Fields are separated
      from each other by the delimiter ")" (ASCII 124).  Fields that do not
      contain data should be zero length or a blank field (empty with no space
      or additional delimiters between the delimiters before and after the
      field) with the delimiter as a place holder.  For the purposes of
      Section 9 of this exhibit, wherever "blank" is given as an option under
      the "Format/Contents" column, it refers to a blank field as explained
      above.

1.3   Numeric fields may contain numeric digits, a decimal place, and a
      leading minus sign.  A positive sign is assumed if no negative sign is
      entered in a numeric field and shall not be entered into any numeric
      field.  Values that exceed the maximum length allowed shall be reported
      to the maximum possible, maintaining the specified decimal place and
      maximum field length restrictions.

1.4   Requirements for significant figures and number of decimal places are
      specified in Exhibit B.  The numeric field lengths are specified such
      that all possible numeric values can be written to the file.  The size
      of the numeric field indicates the maximum number of digits, including a
      decimal place and negative sign (if appropriate), that can appear in the
      field at the same time.  Therefore, the number reported may need to be
      rounded (using rounding rules described in Exhibit B)  to fit into the
      field.  The rounding shall maintain the greatest significance possible
      providing the field length limitation.   In addition, the rounded number
      that appears on the form,  and therefore in the field on the diskette
      file, must be used in any calculation that may result in other numbers
      reported on the same form or other forms in the SDG.  The numbers/values
      reported by the Contractor are used by CCS to calculate a result (e.g.,
      CRQL).  The final value calculated by CCS is then rounded according to
      rounding rules described in Exhibit B and is used for comparison to the
      final value (e.g., CRQL) reported by the Contractor.  Field lengths
      should only be as long as necessary to contain the data; packing with
      blanks is not allowed.
                                      H-3                               OLM03.0

-------
Exhibit H — Section 2
Record Types
2.0   RECORD TYPES

2.1   The Agency Standard consists of variable length ASCII records.  Maximum
      field length specifications match the reporting requirements in Exhibit
      B.  The last two bytes of each record shall contain "carriage return"
      and "line feed", respectively.

2.2   This implementation consists of twelve record types that can be
      summarized in four groups, designated by the first record type in each
      group:

       Type              Type ID   Contents

       Run Header           10      Information pertinent to a group of
                                   samples processed in a continuous
                                   sequence; usually several per SDG
       Sample Header        20      Sample identifying, qualifying, and
                                   linking information
       Results Record       30      Analyte results and qualifications
       Comments Record      90      Free form comments
2.3   A separate run header is used for volatiles (VOA),  semivolatiles (SV),
      and for each column analysis for pesticides (PEST)  (minimum of four type
      10 series for VOA/SV/PEST SDG).   The 20 series records contain sample
      character'sties and link samples within an SDG to  the corresponding
      calibrations, blanks, and other QCs.  The 30 series records contain the
      actual analytical results by analyte within each sample.   The 10,  20,
      and 30 records are associated with each other by their position in the
      file (i.e.,  30 series records follow the corresponding 20 series,  which
      in turn follow the 10 series run header records).
                                     H-4                               OLM03.0

-------
                                                        Exhibit H — Section 3
                                                               Production Runs
3.0   PRODUCTION RUNS

3.1   A production run represents a "group" or "batch" of samples that are
      processed in a continuous sequence under relatively stable conditions.
      Specifically:

3.1.1    Calibration - All samples in a run use the same initial calibration
         data.

3.1.2    Method number - Constant throughout a run.

3.1.3    Instrument conditions - Constant throughout a run.

3.2   Each instrumental analysis consists of a separate production run and is
      reported in a separate file.  There will be a separate production run
      for each of the two pesticide GC columns utilized.   Thus,  a full three
      fraction analysis will consist of a minimum of four production runs.
                                     H-5                               OLM03.0

-------
Exhibit H — Section 3
Production Runs
3.3   Example of the Sequence of Record Types in a File1
   10              Contains Run Header information.
       11          Contains additional run-wide information.
       20          Occurs once for each sample, calibration,  mean response
                   factor,  matrix spike duplicate result,  etc.   Acts as a
                   header.
       21
       22          Contains additional information for samples.
       23
       27
           30      Occurs once for each final analytical result.  Reports
                   the value being determined as defined by the  type 20.
               32  Reports  any auxiliary data necessary.
               33  Reports  compound names for tentatively identified
                   compounds (TICs)  if necessary.
               36  Reports  any instrumental data necessary.
           30      Values for the next analyte or parameter being measured.
               32  f.dditional data may vary for each parameter,  and may
               33  occur in any order.  Multiple occurrences  of  the same
               36  record type, however,  must be consecutive.
           30      Continues for as many as are necessary.
               32
               33
               36
           30
               32
               33
               36
       20          Next Sample Header record.  The following applies to the
       21          next sample or other group of data.
       22
           30
               32
               33
               36
           30
               32
               33
               36
                   etc.
       20
       21
           30
               32
               33
               36
                   etc.
      Appendix A  provides  a  detailed  set of  examples  for the  use  of  the
      different record types,  and their relationship to other record types.

                                     H-6                               OLM03.0

-------
                                                        Exhibit H — Section 4
                                                               Record Sequence
4.0   RECORD SEQUENCE

4.1   The sequence of records for Agency Standard files is as follows:  A Run
      Header (type 10) record shall be present once and once only  (per file.)
      as the first record in a file.  Therefore, a complete VOA/SV/PEST SDG
      will consist of several files.

4.2   Each environmental sample, calibration standard, or quality  control
      sample is represented by a group composed of type 20, 21, 22, 23, and 27
      records,  that hold sample level identifying information, followed by
      type 30,  32, 33, and 36 records for each method analyte including
      surrogates, system monitoring compounds, and internal standards in the
      sample.  The type 20 record holds a count for the number of method
      analytes being determined and includes all target compounds, surrogates,
      system monitoring compounds, and internal standards plus each peak of
      the multi-component pesticides (do not include TICs in this  count).  A
      separate field on the type 23 record contains the number of  TICs found.
      Type 20 records shall occur in the order of sample analysis.  In
      addition, a type 20 record with a QC code "MNC", followed by a type 30
      record for each method analyte {reporting values such as mean response
      factors)  will appear after the type 10 or type 11 record and before the
      type 20 record that initiates the analytical sequence.  Similarly, for
      pesticide runs, a type 20 record with a QC code "GPC" for GPC recovery,
      followed by type 30 records for each of the method analytes  spiked; and
      a type 20 record with a QC code "FLO" for Florisil recovery, followed by
      ty;3 30 records for each of the method analytes (and the two surrogates)
      included . \ the Florisil check will appear before the type 20 record
      that initiates the analytical sequence.

4.3   Type 90 comment records may be defined to occupy any position after the
      type 10 (header) record.
                                      H-7                               OLM03.0

-------
Exhibit H — Sections 5-7
File/Record Integrity/Dates and Times/Multiple Volume Data
5.0   FILE/RECORD INTEGRITY

      All record types shall contain the following check fields to ensure file
      and record integrity:

      Record     Field   Field
      Position   Length  Contents                 Remarks

      First Field    2    Record  type              "10" or as appropriate
      Last Field     5    Record  sequence  number   00001-99999,  numbered
                                                 within file sequentially
                     4    Record  checksum1         Four hexadecimal digits
                     2    Must contain CR  and LF
6.0   DATES AND TIMES

      Date or time-of-day information consists of successive groups of two
      decimal digits, each separated by delimiters.  Dates are given in the
      order YY MM DD, and times as HH MM.  All hours shall be given as 00 to
      23 using a 24-hour clock and shall be local time.   All days shall be
      given as 01 to 31.  All months shall be given as 01 to 12 (e.g., 01 is
      January, 02 is February).

7.0   MULTIPLE VOLUME DATA

      There is no requirement under this format that all the data from an
      entire sample delivery group fit onto a single diskette.  However, each
      single production run must fit onto a single diskette if possible.  If
      that is not possible, then it is necessary that all files start with a
      type 10 record, and that the multiple type 10 records for each file of
      the same production run be identical.  Information for a single sample
      shall not be split between files.
      The checksum is the  sum  of  the ASCII representation of the  data on the
      record up to the Record Sequence Number (not  including the Record Sequence
      Number) plus the checksum of the previous record.   The sum is taken modulo
      65536  (216)  and  is  represented  as  four  hexadecimal  digits  (i.e.,  the
      remainder of the  sum divided  by  65536  represented as  four  hexadecimal
      digits).

                                     H-8                               OLM03.0

-------
                                                         Exhibit  H  —  Section  8
                                                                    Deliverable
 8.0    DELIVERABLE
 8.1    The file  shall  be  submitted  on 5-1/4  inch  floppy  diskette(s), which  may
       be either double-sided,  double density,  360  K-byte  or  high  density
       1.2 M-byte diskette(s).   IBM-compatible, 3.5 inch double-sided,  double
       density 720 K-byte or   high  density 1.44 M-byte diskettes may also be
       submitted.   The diskettes shall be formatted and  recorded using  MS-DOS
       Operating System.   The  diskettes shall  contain all  information relevant
       to one and only one SDG.

 8.2    Agency Standard data from an entire SDG may  not fit onto a  single
       diskette.   If a single  production run is being split onto multiple
       diskettes,  then all files shall start with a type 10 record, and the
       multiple  type 10 records  for each file  of the same  production run shall
       be identical.   Do  not split  the data  from a  single  sample onto multiple
       diskettes.

 8.3    Information on  the diskette  must correspond  to information  submitted in
       the hardcopy raw data package and on the hardcopy raw  data  package
       forms.  For example, type 30 results  field specifies maximum length  of
       13.  When  reporting CRQLs or results on Form 1, maximum length is 13 as
       is  specified in  this exhibit;  when reporting 'calculated amounts' on
       Form 7D,  hardcopy  specified  maximum length is 8.  Unused records shall
       not be included  on  the diskettes.  If the information  submitted  in the
       hardcopy  data package forms  is  changed, the  information in  the diskette
       file shall  be changed accordingly, and  a complete diskette  containing
       all the information for the  SDG  shall be resubmitted along  with  the
       hardcopy  at  no  additional  cost  to the EPA.

 8.4    Each diskette shall  be identified with  an external  label containing  (in
       this order)  the  following  information:

       Disk Density
       File Name(s)
       Laboratory  Name  (optional)
       Laboratory  Code
       Contract Number
       Case Number/SDG
       SAS Number  (where applicable)
       Initial Submission  or Resubmission (as applicable)  and Date

8.5    The format  for File Name  shall be XXXXX.O01  to XXXXX.O99.   Where XXXXX
       is the SDG  identifier,  O designates Organics, and 01 through 99  is the
       file number.

8.6   Dimensions of the label must  be  in the range of 2-1/2" to 2-3/4" long by
       1-1/4" to 1-1/2" wide for a  5-1/4 inch floppy diskette, and 2-1/2" to 2-
       3/4" long by 2"  to 2-1/8" wide for a 3-1/2  inch IBM-compatible diskette.

8.7   Section 9.0  (Record Listing)   provides information for the usage of each
      of the record types.  Where  specified, labels indicate the  nature of the
      value(s)  that follow on that  record.   If the value(s)  will  not be
      reported,  the label shall be  omitted.
                                      H-9                               OLM03.0

-------
Exhibit H — Section 8
Deliverable
8.7.1    A record type 30 for each TCL compound,  surrogate, system monitoring
         compound and internal standard quantitated for shall be reported.  If
         the TCL is not detected, the 'U' qualifier in the appropriate field
         shall be indicative of that.

8.7.3    For multicomponent analytes (Aroclors/toxaphene), if the
         multicomponent analyte is detected,  a record type 30 and 32 shall be
         reported for each peak identified.
                                     H-10                               OLM03.0

-------
                                                        Exhibit H — Section 9
                                                                Record Listing
9.0   RECORD LISTING

      The following lists every record type required to report data from a
      single SDG.

9,1   Production Run Header Record (Type 10)

      Use:      Each production run will start with a record type 10.

       MAXIMUM
       LENGTH    CONTENTS                      FORMAT/CONTENTS
         2       RECORD TYPE                   "10"
         6       Delimiters                    I I I I I I
         5       INSTRUMENT/DETECTOR           Character1
         1       Delimiter                     |
         8       METHOD NUMBER                 Character2
         2       Delimiters                    ||
         6       LAB CODE                      Character
         4       Delimiters                    Ml!
        11       CONTRACT NUMBER               Character
         1       Delimiter                     |
        10       INSTRUMENT ID                 Character
         2       Delimiters                    ||
        25       LABORATORY NAME               Character
         2       Delimiters                    ||
         5       RECORD SEQUENCE NUMBER        Numeric
         4       CHECKSUM                      Character
      General descriptor  (GC/MS  for  VOA/SVOA  analysis  or  GC for  pesticide
      analysis on GC/EC).

      OLM03.0V  For  Volatiles;  OLM03.0B  for   semivolatiles;  OLM03.0P  for
      pesticides.    (O  for  Organic, L  for  Low,  M  for  Medium,  zero  three  for
      document number,  zero V for volatiles, zero  B  for  semivolatiles,  zero P
      for pesticides.)

                                     H-ll                              OLM03.0

-------
Exhibit H — Section 9
Record Listing
9.2   Chromatography Record (Type 11)

      Use:      To describe chromatograph condition.  Must be present once for
                each production run immediately following the record type 10.

       MAXIMUM
       LENGTH    CONTENTS                      FORMAT/CONTENTS
          2      RECORD TYPE                   "11"
          1      Delimiter                     |
         10      GC COLUMN IDENTIFICATION      Character
          2      Delimiters                    ||
          4      GC COLUMN ID1                  Numeric (mm)
         11      Delimiters                    I I I II I I I I I I
          5      RECORD SEQUENCE NO.            Numeric
          4      CHECKSUM                      Character
      Internal Diameter of the GC column used.

                                     H-12                              OLM03.0

-------
                                                        Exhibit H — Section 9
                                                                Record Listing
9.3   Sample Header Data Record (Type 20)

       MAXIMUM
       LENGTH    CONTENTS
          2      RECORD TYPE
          2      Delimiters
         12      EPA SAMPLE NUMBER

          1      Delimiter
          1      MATRIX
          1      Delimiter
          3      QC CODE
          1      Delimiter
          3      SAMPLE QUALIFIER
          1      Delimiter
          5      CASE NUMBER
          1      Delimiter
          6      SDG NO.
          1      Delimiter
          2      SAMPLE/BLANK/STANDARDS YEAR ANALYZED
          1      Delimiter
          2      SAMPLE/BLANK/STANDARDS MONTH ANALYZED
          1      Delimiter
          2      SAMPLE/BLANK/STANDARDS DAY  ANALYZED
          1      Delimiter
          2      SAMPLE/BLANK/STANDARDS HOUR ANALYZED
          1      Delimiter
          2      SAMPLE/BLANK/STANDARDS MINUTE ANALYZED
          2      Delimiters
          2      SAMPLE WT/VOL  UNITS
          1      Delimiter
          5  .    SAMPLE WT/VOL
                             FORMAT/CONTENTS
                             "20''
                             II
                             As  is exactly  on  the
                             hardcopy  form

                             CHARACTER1
                             I
                             Character (See Section  10)
                             I
                             RIN/REX/REJ/SRN/blank2

                             Numeric
                             I
                             Character
                             I
                             YY
                             I
                             MM

                             DD

                             HH
                             !
                             MM
                             II
                             "G"/"ML"/blank3
                             I
                             Numeric
      "0"  if  not applicable
      soil.
[calibration,  tune,  etc.);  "1" for water; "H"  for
      "RIN"  for  reinjection;   "REX"  for  re-extractions;   "REJ"  for  rejected
      samples;  "SRN" for  dilutions;  and  leave  blank  (empty  field with  zero
      length) when  none  of  the  previous  conditions  apply.   In  case  of  multiple
      operations  on a  sample,  the final  operation will  be  indicated  (e.g.,
      reinjection of a dilution; AAA12DLRE would  have  a QC Code of  "RIN").

      Sample  WT/VOL unit  is  mL (milliliters)  for  liquids  and G  (grams)  for
      solids.   The  sample  units code  indicates which units  are  in  use for  the
      current sample.  Leave blank  (zero length)  if not applicable.

      Sample WT/VOL  is the volume in milliliters for liquid or the wet weight in
      grams for solids.  Sample WT/VOL includes the purge  volume.
                                     H-13
                                          OLM03.0

-------
Exhibit H — Section 9
Record Listing
       Sample Header Data Record (Type 20)  (Cont.)
                 CONTENTS                                FORMAT/CONTENTS
                 Delimiter                               |
                 ANALYTE COUNT                           Numeric5
                 Delimiters                              |||
                 RECORD SEQUENCE NO.                      Numeric
                 CHECKSUM                                Character
     1-3 decimal  digits.   Counts TCL  analytes,  surrogates,  system monitoring
     compounds  (SMC),  internal  standards,  and all  peaks  reported  for multi-
     component  PCBs.   Do  not  include  the count  for TICs  in  this field.   For
     calibrations,  also count DFTPP,  if included in calibration solution.

                                    H-14                              OLM03.0

-------
                                                        Exhibit H — Section 9
                                                                Record Listing
9.4   Sample Header Data Record (Type 21)

      Use:      Continuation of Type 20.
      Position: Follows the Type 20 to which it applies.
                                           FORMAT/CONTENTS
                                           "21"
MAXIMUM
LENGTH
2
1
1
1
1
2
1
2
6
1
14
1
2
1
2
1
2
2
2
1
2
1
2
2
8

CONTENTS
RECORD TYPE
Delimiter
PURGE
Delimiter
LEVEL
Delimiters
EXTRACTION
Delimiters
SAS NUMBER
Delimiter
LAB FILE/SAMPLE ID
Delimiter
YEAR EXTRACTED
Delimiter
MONTH EXTRACTED
Delimiter
DAY EXTRACTED
Delimiters
YEAR RECEIVED
Delimiter
MONTH RECEIVED
Delimiter
DAY RECEIVED
Delimiters
INJECTION/ALIQUOT '
                                           "N" for not heated; "Y" for heated;
                                           blank if SV or PEST
                                           I
                                           "L"/"M"/blank1
                                           II
                                           S/C/N/blank (for volatiles)2

                                           Character

                                           Character

                                           YY/blank (for volatiles)
                                           I
                                           MM/blank (for volatiles)
                                           I
                                           DD/blank (for volatiles)
                                           II
                                           YY/blank (for standards,  tunes and
                                           blanks)
                                           I
                                           MM/blank (for standards,  tunes,  and
                                           blanks)
                                           1
                                           DD/blank (for standards,  tunes,  and
                                           blanks)
      "L"  for low level samples and  "M" for medium  level  samples  for volatile
      and  semivolatile analyses.   Leave blank  for pesticides,  all  calibrations
      and  all tunes.

      "S"  for separatory funnel; "C" for continuous liq-lig; "N" for sonication;
      blank (zero length field)  for volatiles.

      Lab  File ID for  volatile  and semivolatile analyses.  Lab Sample  ID for
      pesticides  in  same format as on forms.

      Injection volume, in  uL,  for  SVOAs  and  PESTs;  Soil Aliquot  Volume for
      medium level VOA.
                                     H-15
OLM03.0

-------
Exhibit H — Section 9
Record Listing
       Sample Header Data Record (Type 21)  (Cont.)
       MAXIMUM
       LENGTH
          2
          5
          4
CONTENTS
Delimiters
RECORD SEQUENCE NO.
CHECKSUM
FORMAT/CONTENTS
II
Numeric
Character
9.5   Sample Condition Record (Type 22)

      Use:       Continuation of type 20.  Used to describe additional Sample
                Conditions.
      Position:  Follows the  type 20 and  21 to which it applies.
                                           FORMAT/CONTENTS
                                           "22"

                                           YY/blank (for PEST)1

                                           MM/blank (for PEST)

                                           DD/blank (for PEST)

                                           HH/blank (for PEST)

                                           MM/blank (for PEST)

                                           Character/blank (for PEST)2
                                           I
                                           Numeric/blank (for aqueous samples
                                           and volatiles)
                                           I
                                           Numeric
                                           I
                                           "Y"/"N"/blank (for volatiles)

                                           Numeric/blank (for low level VOA)3
      For volatiles and semivolatiles, enter the date  and  time of analysis of
      the most recent  50 ug/L  (VOAs) or the 50 ng (SVGAs) standard run prior to
      the sample reported in  the  associated  type 20 record.   Leave  blank for
      pesticides.

      Lab File  ID  of  standard  specified  in  1  above  (volatiles/semivolatiles
      only) .    This field  must match the Lab  File  ID on Type  21  for  the
      associated calibration (VSTD050/SSTD050).   Leave blank for pesticides.
MAXIMUM
LENGTH
2
1
2
1
2
1
2
1
2
1
2
1
14
1
4
1
5
1
1
1
8

CONTENTS
RECORD TYPE
Delimiter
CALIBRATION YEAR
Delimiter
CALIBRATION MONTH
Delimiter
CALIBRATION DAY
Delimiter
CALIBRATION HOUR
Delimiter
CALIBRATION MINUTE
Delimiter
CALIBRATION FILE ID
Delimiter
PH
Delimiter
PERCENT MOISTURE
Delimiter
DECANTED
Delimiter
EXTRACT VOLUME
                                     H-16
                                                      OLM03.0

-------
                                                  Exhibit H — Section 9
                                                          Record Listing
 Sample Header Data Record (Type 22) (Cont.)

 MAXIMUM
 LENGTH    CONTENTS                  FORMAT/CONTENTS
    1      Delimiter                 |
    8      DILUTION FACTOR           Numeric1
    3      Delimiters                | |  |
    5      LEVEL                     Numeric/blank (for VOA/SV)5
    1      Delimiter                 |
    5      RECORD SEQUENCE NO.       Numeric
    4      CHECKSUM                  Character
(...continued)
Enter the  Soil Extract  Volume for medium  level VOA,  and Concentrated
Extract Volume  for  all  SVGA and PEST.  The  value  should be reported in
microliters.

Dilution factor of sample analyzed (omit contract-mandated dilutions).

Concentration  level of  Pesticide  Individual Mix  A  and  B  standards.
Concentration of low point, mid point and high point calibration standards
as a multiplier of  low point.  Low  point  =  1.0; Mid  point = 4.0; High
point > 16.0.
                               H-17                              OLM03.0

-------
Exhibit H — Section 9
Record Listing
9.6   Associated Injection and Counter Record (Type 23)

      Use:      Continuation of type 20.  Used to identify associated blanks
                and tunes, and the number of surrogates/SMCs and spikes
                outside of the QC limits  and the number  of TICs.
      Position: Follows the type 20, 21,  and 22  to which it applies.
                 CONTENTS                 FORMAT/CONTENTS
                 RECORD TYPE              "23"
                 Delimiter                |
                 INSTRUMENT PERFORMANCE   "P"  (for BFB and DFTPP IPC)  or blank (for
                 CHECK (IPC/TUNE)  LABEL   pesticides)
          1      Delimiter                |
          2      IPC/TUNE INJECTION YEAR  YY/blank (for PEST)
          1      Delimiter                |
          2      IPC/TUNE INJECTION       MM/blank (for PEST)
                 MONTH
          1      Delimiter                |
          2      IPC/TUNE INJECTION DAY   DD/blank (for PEST)
          1      Delimiter                |
          2      IPC/TUNE INJECTION HOUR  HH/blank (for PEST)
          1      Delimiter                ]
          2      IPC/TUNE INJECTION       MM/blank (for PEST)
                 MINUTE
          1      Delimiter                |
         14      DFTPP/BFB LAB FILE ID     Character/blank (for PEST)
          1      Delimiter                |
          2      VOLATILE STORAGE  BLANK   "HB"  (for VOA)  or blank (for SV and PEST)
                 LABEL
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  YY/blank (for SV and PEST)
                 YEAR
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  MM/blank (for SV and PEST)
                 MONTH
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  DD/blank (for SV and PEST)
                 DAY
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  HH/blank (for SV and PEST)
                 HOUR
          1      Delimiter                |
          2      STORAGE BLANK INJECTION  MM/blank (for SV and PEST)
                 MINUTE
          1      Delimiter                |
         14      STORAGE BLANK LAB        Character
                 FILE ID (VOA ONLY)
          4      Delimiters               II I I
                                     H-18                               OLM03.0

-------
                                                  Exhibit H — Section 9
                                                          Record Listing
 Associated  Injection  and Counter Record  (Type  23)  (Cont.)
MAXIMUM
LENGTH
2
1
2

1
2

1
2

1
2

1
2

1
14



1
1


1
2


1
1
1
2
1
1
1
2

CONTENTS
METHOD BLANK LABEL
Delimiter
METHOD BLANK INJECTION
YEAR
Delimiter
METHOD BLANK INJECTION
MONTH
Delimiter
METHOD BLANK INJECTION
DAY
Delimiter
METHOD BLANK INJECTION
HOUR
Delimiter
METHOD BLANK INJECTION
MINUTES
Delimiter
METHOD BLANK LAB
FILE (for VGA and
SV) /SAMPLE ID (for
PEST)
Delimiter
SURROGATE (for SV and
PEST) /SMC (for VOA)
RECOVERY LABEL
Delimiter
SURROGATE (for SV and
PEST) /SMC (for VOA)
RECOVERIES OUT
Delimiter
TIC LABEL
Delimiter
NO. OF TICS
Delimiter
SPIKE RECOVERY LABEL
Delimiter
SPIKE RECOVERIES OUT
                                    FORMAT/CONTENTS
                                    "MB"/blank  (for standard, tune and method
                                    blanks)
                                    I
                                    YY/blank  (for standard, tune and method
                                    blanks)
                                    I
                                    MM/blank  (for standard, tune and method
                                    blanks)

                                    DD/blank  (for standard, tune and method
                                    blanks)

                                    HH/blank  (for standard, tune and method
                                    blanks)
                                    I
                                    MM/blank  (for standard, tune and method
                                    blanks)

                                    CHARACTER
                                    "P" for % recoveries/blank  (for STD/IPC)
                                    Numeric
                                    "T" (for VOA and SV TICs)/blank  (for
                                    PEST)

                                    Numeric

                                    "S" for Matrix Spikes and Matrix Spike
                                    Duplicates/blank for anything else
                                    I
                                    Numeric/blank2
This will be the  number  of surrogate (for SV or  PEST)  or SMC (for VOA)
recoveries outside  QC  limits for a  specific column.    It  should not be
cumulative of the two columns for pesticides.
                               H-19
OLM03.0

-------
Exhibit H — Section 9
Record Listing
       Associated Injection and Counter Record (Type 23)  (Cont.)

       MAXIMUM
       LENGTH    CONTENTS                 FORMAT/CONTENTS
          1      Delimiter                |
          1      RPD LABEL                "R"  for RPD/blank3
          1      Delimiter                |
          2      RPD OUT                  Numeric
          1      Delimiter                |
          5      RECORD SEQUENCE NO.       Numeric
          4      CHECKSUM                 Character
     (...continued)
     Enter the number of spike recoveries out.   Enter "0"(zero)  if none of  the
     spike recoveries are outside of the QC limit.

     "R"  for Matrix Spike/Matrix  Spike Duplicate  Recovery Relative  Percent
     Differences.  Leave blank for all  other  samples (only  report  for MS/MSD).
                                    H-20                               OLM03.0

-------
                                                        Exhibit H — Section 9
                                                                Record Listing
9.7   Sample Cleanup Record (Type 27)

      Use:      Continuation of type 20. Used to identify sample/blank cleanup
                procedures and QC results.
      Position: Follows type 20, 21, 22, and 23 to which it applies.
                                                   FORMAT/CONTENTS
                                                   "27"


                                                   "G" for GPC/blank (for VOA)1

                                                   I
                                                   YY/blank (for VOA)


                                                   MM/blank (for VOA)


                                                   DD/blank (for VOA)


                                                   HH/blank (for VOA)
                                                   I
                                                   I
                                                   MM/blank (for VOA)
                                                   Character/blank (for VOA and
                                                   SV)2
                                                   "F"  (for PEST)  or blank (for
                                                   VOA  and SV)
                                                   I
                                                   YY/blank (for VOA and SV)

                                                   MM/blank (for VOA and SV)

                                                   DD/blank (for VOA and SV)

                                                   HH/blank (for VOA and SV)

                                                   MM/blank (for VOA and SV)

                                                   Character
      "G"  indicates that GPC was performed.   If GPC was not performed, leave the
      field  blank.

      Lab  Sample  ID of associated GPC.  This is a unique identifier assigned to
      the  spike  recovery  results  for  a specific  GPC  calibration check  for
      pesticides.  Leave blank  for  volatiles and semivolatiles.
MAXIMUM
LENGTH
2
1
1
1
2
1
2
1
2
1
2
1
2
1
14
1
1
1
2
1
2
1
2
1
2
1
2
1
14

CONTENTS
RECORD TYPE
Delimiter
FIRST CLEANUP TYPE
Delimiter
GPC CALIBRATION CHECK YEAR
Delimiter
GPC CALIBRATION CHECK MONTH
Delimiter
GPC CALIBRATION CHECK DAY
Delimiter
GPC CALIBRATION CHECK HOUR
Delimiter
GPC CALIBRATION CHECK MINUTE
Delimiter
GPC Data Descriptor
Delimiter
FLORISIL CLEANUP TYPE
Delimiter
FLORISIL LOT CHECK YEAR
Delimiter
FLORISIL LOT CHECK MONTH
Delimiter
FLORISIL LOT CHECK DAY
Delimiter
FLORISIL LOT CHECK HOUR
Delimiter
FLORISIL LOT CHECK MINUTE
Delimiter
FLORISIL DATA DESCRIPTOR
      Lab  Sample  ID  of  associate  Florisil  lot  check.
      identifier  assigned to  a  lot  of  Florisil cartridges.
This  is  a  unique
                                     H-21
                                                                       OLM03.0

-------
Exhibit H — Section 9
Record Listing
       Sample Cleanup Record (Type 27) (Cont.)

       MAXIMUM
       LENGTH    CONTENTS
          1      Delimiter
          1      SULFER CLEANUP

          1      Delimiter
          2      SULFUR BLANK LABEL
          1      Delimiter
          2      SULFER BLANK INJECTION YEAR
          1      Delimiter
          2      SULFER BLANK INJECTION MONTH
          2      Delimiters
          2      SULFER BLANK INJECTION DAY
          1      Delimiter
          2      SULFER BLANK INJECTION HOUR
          1      Delimiter
          2      SULFER BLANK INJECTION MINUTE
          1      Delimiter
         14      SULFUR BLANK LABORATORY/
                 SAMPLE ID
          1      Delimiter
          5      RECORD SEQUENCE NO.
          4      CHECKSUM
FORMAT/CONTENTS

Y/N (for PEST)/blank  (for VOA
and SV)

"SB"/blank  (if no separate
sulfur blank was prepared for
pesticides; also blank for VOA
and SV)
I
YY/blank (for VOA and SV)
I
MM/blank (for VOA and SV)
II
DD/blank (for VOA and SV)
I
HH/blank (for VOA and SV)

MM/blank (for VOA and SV)

Character
Numeric
Character
                                     H-22
                    OLM03.0

-------
9.8   Results Data Record  (Type  30)
                                                        Exhibit H  — Section 9
                                                                Record Listing
          1
          9
          1
          5
          1
          3
          1
         13
          1
          5
          1
          1
          1
         13
          1
                 CONTENTS
                 RECORD TYPE
                 Delimiter
                 ANALYTE LABEL

                 Delimiter
                 CAS NUMBER
Delimiter
INTERNAL STD. CAS NUMBER
Delimiter
CONCENTRATION UNITS
Delimiter
RESULT QUALIFIER
Delimiter
RESULTS
Delimiter
FLAGS
Delimiter
AMOUNT ADDED LABEL
Delimiter
AMOUNT ADDED
Delimiter
                              FORMAT/CONTENTS
                              "30"
                              I
                              "C" for CAS Number  (blank for
                              unknown TICs)
Numeric  (for TCL, surrogates,
DFTPP, BFB, SMC, internal
standards and identified TICs)
I
Numeric

Character "ug/L" (aqueous);
"ug/Kg"  (soil); "ng"  (amount
added)

Character1'
I
Numeric

Character4
I
"A" for Amt. added5
i
Numeric
1     When  a  Type  20  Record  is  used  for  calibration  summary  (MNC),  the
      associated Type 30 Record uses "AVG" for average RRFs  and Mean Calibration
      Factors.  See Exhibit H Section 10.3.2.

2     For pesticide sample analysis,  if  an  analyte is detected in only one of
      the two  column analyses,  report the  analyte as "not  detected"  in both
      runs.    Report  result qualifier, for  each  column,  as  BDL.   See Section
      10.3.2 for result qualifiers.

3     Leave this field blank for reporting of all non-detects.

4     A maximum of five flags (D,E,J,B,A,P,C,X,Y,Z or N)  with no space between
      the flags can be reported, each representing a qualification of the result
      as described in  Exhibit B.  For  surrogates,  the 'D'  flag  will  indicate
      surrogates diluted out.

5     For Matrix Spike/Matrix Spike Duplicate analysis, surrogate, SMC for VOA,
      SV and PEST (Form  3s).  Nominal Amount for Pesticides  (Form  7D/7E).  Spike
      added for florisil and GPC (Form 9A/9B).
                                     H-23
                                                      OLM03.0

-------
Exhibit H — Section 9
Record Listing
       Results Data Record (Type 30)  (Cont.)
       MAXIMUM
       LENGTH    CONTENTS
          1      CRQL LABEL

          1      Delimiter
         13      CRQL
          1      Delimiter
          1      RSD LABEL
          1      Delimiter
          5      RSD VALUE
          1      Delimiter
          1      MS/MSD  REC LABEL
          1       Delimiter
          5       MS  %  RECOVERY

          1       Delimiter
          5       MSD % RECOVERY

          1       Delimiter
          1       RPD LABEL
          1      Delimiter
          5      RPD VALUE
          1      Delimiter
          1      SURR/SPIKE RECOVERY LABEL
         1      Delimiter
         5      SURR/SPIKE RECOVERY
FORMAT/CONTENTS
"U" for  "undetected" or  blank  when
analyte  is detected
I
Numeric
I
"R" for  % Resolution/RSD6
I
Numeric
I
"P" for  % recovery [MS/MSD]/blank
(for sample [except MS/MSD],
standard, tune, blanks,
calibration)

Numeric/blank  (for everything
except MS)
I
Numeric/blank  (for everything
except MSD)

"D" for MS/MSD or for pesticide
calibration verification
(%D)/blank
I
Numeric/blank7

"S" for % recovery/blank (for non-
surrogate/SMC and non-spike
analytes

%  Recovery/blank8
     "R"  for % Resolution  (Form 6G)  or for RSD  of Response  factors under
     Calibration summary  (MNC) Type  20.   "R"  for %Resolution is optional  for
     the PEM (6H);  Individual Standard Mixture A  (61); and Individual Standard
     Mixture B (6J) in the electronic deliverable only.   (Blank  for VOA and SV
     fractions.)
     RPD for MS/MSD recoveries, or %D for pesticides.
     (Form 7D/7E).  Otherwise, leave blank.
       Calibration Verification
     Surrogate (for SV and PEST)/SMC (for VOA) or Spike (Forms 2, Form 9A/9B)
     recovery.  Leave blank for non-surrogate and non-spike analytes.
                                    H-24
                                                                      OLM03.0

-------
       Results Data Record  (Type  30)  (Cont.)
                                                        Exhibit  H  —  Section  9
                                                                 Record  Listing
       MAXIMUM
       LENGTH    CONTENTS
           1      Delimiter
FORMAT/CONTENTS
                 MEAN CONCENTRATION LABEL
          1      Delimiter
         13      MEAN CONCENTRATION

          1      Delimiter
          1      PERCENT DIFFERENCE LABEL

          1      Delimiter
          5      PERCENT DIFFERENCE
          1      Delimiter
          1      INTERNAL STANDARD AREA LABEL

          1      Delimiter
         13      INTERNAL STANDARD AREA

          1      Delimiter
          5      RECORD SEQUENCE NO.
          4      CHECKSUM
"M" for Mean cone,  (for
multicomponent PEST only)/blank
(for VOA and SV)
I
Numeric (for PEST)/blank (for VOA
and SV)9

"F" or "P" (PEST)/blank (for VOA
and SV field sample analysis)
i
Numeric

"I" for IS Area  (for VOA and
SV)/blank (for PEST)
I
Numeric (for VOA and SV)/blank
(for PEST)
I
Numeric
Character
9     Mean  Concentration for  Multicomponent analytes  detected  in pesticide
      analyses.

10    "P"  for  Percent Difference between  concentrations from  two columns in
      pesticide analyses,  or "F" for  Percent Difference between average RRF
      (initial  calibration)  and  RRF50  (continuing  calibration)  in VOA/SVOA
      analyses.  Leave blank for  volatile  and semivolatile sample, blank, and
      tune analysis.
                                     H-25
                        OLM03.0

-------
Exhibit H — Section 9
Record Listing
9.9   Auxiliary Data Record (Type 32)

      Use:      Used to report retention time (in minutes)  for Internal
                Standards and for Tics (for Volatiles and Semivolatiles).  Used
                to report retention time data and percent breakdown (for
                pesticides).
      Position: Follows type 30.  (Record will only be required as specified
                above.)
       MAXIMUM
       LENGTH    CONTENTS
          2      RECORD TYPE
          3      Delimiters
          2      RETENTION TIME LABEL
          1      Delimiter
          5      RETENTION TIME
          1      Delimiter
          3      FIRST LIMIT LABEL
          1      Delimiter
          5      RT WINDOW LOWER LIMIT
          1      Delimiter
          3      SECOND LIMIT LABEL
          1      Delimiter
          5      RT WINDOW UPPER LIMIT
          2      Delimiters
          2      % BREAKDOWN LABEL

          1      Delimiter
          5      % BREAKDOWN
          1      Delimiter
          5   .   COMBINED % BREAKDOWN
          2      Delimiters
          1      PEAK
          1      Delimiter
          5      RECORD SEQUENCE NO.
          4      CHECKSUM
FORMAT/CONTENTS
"32"
III
"RT"

Numeric

"RTF"

Numeric
I
"RTT"
I
Numeric
II
"PB" for % breakdown/blank  (for
VOA and SV)

Numeric (DDT/ENDRIN)/blank  (for
VOA and SV)
Numeric/blank (for VOA and SV)
1 THROUGH 5 (for pesticide
multicomponent compounds)/blank
(for VOA and SV)2
Numeric
Character
      The combined %breakdown will be reported on both the record type 32s (for
      DDT and Endrin).

      For positively identified compounds, a minimum of 3 peaks and a maximum of
      5 peaks are allowed.   Types 30 and 32 will be repeated for each peak that
      is reported (a minimum of three, a maximum of  five times).   This is for
      multicomponent analytes  in  pesticide analyses.
                                     H-26
                        OLM03.0

-------
                                                        Exhibit H — Section 9
                                                                Record Listing
9.10  Name Record (Type 33)

      Use:      This record type is used for volatile and semivolatile
                analyses only to carry an analyte name for TICs.  This record
                is not used for pesticide analysis.
      Position: Follows types 30 and 32 for TICs.

       MAXIMUM
       LENGTH    CONTENTS                      FORMAT/CONTENTS
          2      RECORD TYPE                   "33"
          1      Delimiter                     |
         67      NAME OF COMPOUND              Character
          1      Delimiter                     |
          5      RECORD SEQUENCE NO.           Numeric
          4      CHECKSUM                      Character
                                     H-27                              OLM03.0

-------
Exhibit H — Section 9
Record Listing
9.11  Instrumental Data Readout Record (Type 36)

      Use:      This record type is only used for volatile and semivolatile
                analyses to describe DFTPP/BFB percent abundances.  This
                record is not used for pesticide analysis.
      Position: Follows type 30 for DFTPP/BFB data.
          1
          3
          1
          5

          1
          5
          1
          3
          1
          5

          2
          3
          1
          5

          1
          5
          1
          3
          1
          5

          1
          5
          1
          3
          1
CONTENTS
RECORD TYPE
Delimiter
MASS LABEL
Delimiters
FIRST MASS  (DFTPP/BFB)
Delimiters
FIRST PERCENT RELATIVE
ABUNDANCE
Delimiter
SECOND MASS
Delimiter
SECOND PERCENT RELATIVE
ABUNDANCE
Delimiter
PERCENT MASS OF 69
Delimiter
THIRD MASS
Delimiter
THIRD PERCENT RELATIVE
ABUNDANCE
Delimiters
FOURTH MASS
Delimiter
FOURTH PERCENT RELATIVE
ABUNDANCE
Delimiter
PERCENT MASS OF 69
Delimiter
FIFTH MASS
Delimiter
FIFTH PERCENT RELATIVE
ABUNDANCE
Delimiter
PERCENT MASS OF 174
Delimiter
SIXTH MASS
Delimiter
FORMAT/CONTENTS
"36"

"M"
III
Numeric  (DFTPP for SV or BFB  for VOA)
II
Numeric


Numeric

Numeric


Numeric, DFTPP only/blank  (for VOA)

Numeric
I
Numeric

II
Numeric

Numeric


Numeric, DFTPP only/blank  (for VOA)

Numeric

Numeric


Numeric, BFB only/blank (for  SV)

Numeric
                                     H-28
                                                      OLM03.0

-------
                                                 Exhibit H — Section 9
                                                         Record Listing
Instrumental Data Readout Record (Type 36)  (Cont.)
   2
   3
   1
   5

   1
   5
   1
   3
   1
   5

   1
   5
   1
   3
   1
   5

   1
   5
   1
   3
   1
   5

   2
   3
   1
   5

   2
   3
   1
   5

   2
   3
   2
CONTENTS
SIXTH PERCENT RELATIVE
ABUNDANCE
Delimiters
SEVENTH MASS
Delimiter
SEVENTH PERCENT RELATIVE
ABUNDANCE
Delimiter
PERCENT MASS OF 174
Delimiter
EIGHTH MASS
Delimiter
EIGHTH PERCENT RELATIVE
ABUNDANCE
Delimiter
PERCENT MASS OF 174
Delimiter
NINTH MASS
Delimiter
NINTH PERCENT RELATIVE
ABUNDANCE
Delimiter
PERCENT MASS OF 176
Delimiter
TENTH MASS
Delimiter
TENTH PERCENT RELATIVE
ABUNDANCE
Delimiters
ELEVENTH MASS
Delimiter
ELEVENTH PERCENT
RELATIVE ABUNDANCE
Delimiters
TWELFTH MASS
Delimiter
TWELFTH PERCENT RELATIVE
ABUNDANCE
Delimiters
THIRTEENTH MASS
Delimiters
                                    FORMAT/CONTENTS
                                    Numeric
Numeric

Numeric


Numeric, BFB only/blank (for SV)
I
Numeric

Numeric


Numeric, BFB only/blank (for SV)

Numeric
I
Numeric


Numeric, BFB only/blank (for SV)
i
Numeric/blank (for VOA)

Numeric/blank (for VOA)

I
Numeric/blank (for VOA)

Numeric/blank (for VOA)

I
Numeric/blank (for VOA)

Numeric/blank (for VOA)

II
Numeric/blank (for VOA)
                              H-29
                                                      OLM03.0

-------
Exhibit H — Section 9
Record Listing
       Instrumental Data Readout Record (Type 36) (Cont.)

       MAXIMUM
       LENGTH    CONTENTS                  FORMAT/CONTENTS
          5      THIRTEENTH PERCENT        Numeric/blank (for VOA)
                 RELATIVE ABUNDANCE
          1      Delimiter                 |
          5      PERCENT MASS OF 442       Numeric,  DFTPP only (blank for VOA)
          1      Delimiter                 |
          5      RECORD SEQUENCE NO.       Numeric
          4      CHECKSUM                  Character
                                     H-30                               OLM03.0

-------
                                                        Exhibit H — Section 9
                                                                Record Listing
9.12  Comment Record (Type 90)

      Use:      To provide for operator-entered comments.
      Position: May occur anywhere in the file after the type 10 record.
                 CONTENTS                      FORMAT/CONTENTS
                 RECORD TYPE                   "90"
                 Delimiter                     |
                 ANY COMMENT                   Character
                 Delimiter                     |
                 RECORD SEQUENCE NO.           Numeric
                 CHECKSUM                      Character
                                     H-31                               OLM03.0

-------
Exhibit H — Section 10
Definitions of Various Codes
10.0  DEFINITIONS OF VARIOUS CODES USED IN AGENCY  STANDARD RECORDS

10.1  Quality Control and Related Codes (QCC)  in Type 20 Records

10.1.1   Note:   These codes appear in the QC code  fields of type 20 records.
         They are used to indicate the type of data that is being reported.
         QCC  Name

         LRB  LABORATORY (REAGENT)
              BLANK

         LIB  LABORATORY INSTRUMENT
              BLANK

         LSB  LABORATORY SULFUR BLANK
Definition

The "Method Blank" (see Exhibit G)


The "Instrument Blank".


If different from "Method Blank"
(pesticides).
         LHB  LABORATORY STORAGE BLANK   The storage blank (volatiles).
         LSD  LABORATORY SPIKE
              DUPLICATE BACKGROUND
              (ORIGINAL) VALUES
         LF1  LABORATORY SPIKED SAMPLE
              - FINAL - FIRST MEMBER

         LF2  LABORATORY SPIKED SAMPLE
              - FINAL - SECOND MEMBER
An environmental sample which is
analyzed according to the analytical
method, and subsequently used for
the matrix spike and the matrix
spike duplicate (see Exhibit G).

The "Matrix Spike" (see Exhibit G);
must precede LF2.

The "Matrix Spike Duplicate" (see
Exhibit G).
         LPC  LABORATORY PERFORMANCE
              CHECK SOLUTION
A solution of DFTPP (SVOA) or BFB
(VOA) or method analytes  (PEST/PCB)
used to evaluate the performance of
an instrument with respect to a
defined set of criteria (Tune or
Resolution Check Sample)  (see
Exhibit G).
         FLO  FLORISIL CHECK SOLUTION
         GPC  GPC CHECK SOLUTION
A solution of pesticides used to
check recovery from each lot of
Florisil cartridges.  These recovery
results will  be provided in every
production run where associated
samples are analyzed.

A solution of pesticides used to
check recovery from each new GPC
calibration.  These recovery results
will be provided in every production
run where associated samples are
analyzed.
                                     H-32
                              OLM03.0

-------
                                                       Exhibit H — Section 10
                                                  Definitions of Various Codes
         CLM  INITIAL CALIBRATION -
              MULTI-POINT
         CLS  INITIAL CALIBRATION
              SINGLE POINT


         CLC  CONTINUING CHECK
              CALIBRATION

         CLE  CONTINUING PERFORMANCE
              CHECK
10.1.2
                                 The Initial Calibration for GC/MS
                                 (see Exhibit G),  or the Initial
                                 Individual  Standard Mixes (A,  B)  for
                                 pesticides  (see Exhibit D PEST).
                                 Response  factors  (GC/MS)  or
                                 Calibration Factors (pesticides)
                                 will be reported  on the following
                                 type 30 records.

                                 The Initial Toxaphene/Aroclor  Mixes
                                 used to determine all  calibration
                                 factors  (see Exhibit D PEST).

                                 The continuing  calibration
                                 (VSTD50/SSTD50) for GC/MS.

                                 The subsequent  Individual Standard
                                 Mixes (A,B),  Performance  Evaluation
                                 Mixture,  and for  subsequent
                                 injections  of Toxaphene/Aroclor
                                 mixes for pesticides (see Exhibit D
                                 PEST).

                                 A  calibration solution as above used
                                 both as an  initial  calibration (CLM)
                                 and a continuing  check (CLC).   (50
                                 level initial calibration if needed
                                 for Form  8.)
The following QCC values are used on type 20 records which act as a
header, and indicate that additional (usually calculated) analyte
specific data will be present on type 30 (and following type)
records.  Usually, these data will apply to an entire production run,
in which case they will appear immediately following the type 10
record or type 11 record if present.  If the data apply to only a
portion of the samples in the run, they shall be placed immediately
preceding the samples to which they apply.   Much of the rest of the
information in the type 20 record may be blank, indicating that these
data do not apply to these results.
         CLD  DUAL PURPOSE CALIBRATION
         MNC   MEAN VALUES FROM
               CALIBRATIONS
                              The data following represent mean
                              values and percent RSDs  from the
                              initial calibration  (GC/MS) or the
                              mean calibration factors, mean
                              retention times and  retention time
                              windows (pesticides).
                                     H-33
                                                              OLM03.0

-------
Exhibit H — Section 10
Definitions of Various Codes
10.2  Codes For Sample Medium (Matrix, Sources)

            Medium                                       Code
            All Media, Specific Medium not Applicable.    0  (zero)
            Use for Calibrations, Tunes, etc.
            Water                                            1
            Soil                                             H

10.3  List of Sample and Result Qualifiers

      Definition:  A sample qualifier consists of three characters which act
      as an indicator of the fact and the reason that the subject analysis (a)
      did not produce a numeric result, or (b) produced a numeric result for
      an entire sample but it is qualified in some respect relating to the
      type or validity of the result.

10.3.1   Sample Qualifiers


         Qualifier   Full Name     Definition

         RIN         RE-ANALYZED   The indicated analysis results were
                                   generated from a re-injection of the same
                                   sample extract or aliquot (RE SUFFIX).

         REX         RE-PREPARED   The indicated analysis results were
                                   generated from a re-extraction of the same
                                   sample (RE SUFFIX).

         REJ         REJECTED      The results for the entire sample analysis
                                   have been rejected for an unspecified
                                   reason by the laboratory.  For initial
                                   calibration data, these data were not
                                   utilized in the calculation of the mean.

         SRN         DILUTED       The indicated analysis results were
                                   generated from a dilution of the same
                                   sample (DL SUFFIX).

10.3.2   Result Qualifiers in Type 30 Records

         A result qualifier consists of three characters which act as an
         indicator of the fact and the reason that the subject analysis (a)
         did not produce a numeric result,  or (b) produced a numeric result
         for a single analyte but it is qualified in some respect relating to
         the type or validity of the result.   This qualifier is complementary
         to the flags field on a type 30 record.   A TIC must have either a
         TIE,  TFB, ALC,  or PRE result qualifier.
         BDL  BELOW DETECTABLE        Indicates compound was analyzed for but
              LIMITS                  not detected (Form 1 "U" Flag).

         NAR  NO ANALYSIS RESULT      There is no analysis result required
                                      for this subject parameter.

                                     H-34                              OLM03.0

-------
                                              Exhibit H — Section 10
                                         Definitions of Various Codes
AVG  AVERAGE VALUE
CBC  CANNOT BE CALCULATED
LTL  LESS THAN LOWER
     CALIBRATION LIMIT
GTL  GREATER THAN UPPER
     CALIBRATION LIMIT
LLS  LESS THAN LOWER
     STANDARD
TIE  TENTATIVELY IDENTIFIED
     ESTIMATED VALUE
REJ  REJECTED


STD  INTERNAL STANDARD


STB  INTERNAL STANDARD
     BELOW DETECTION LIMITS

FBK  FOUND IN BLANK
TFB  TENTATIVELY IDENTIFIED
     AND FOUND IN BLANK

ALC  ALDOL CONDENSATION
NRP  NON-REPRODUCIBLE
Average value — used to report a range
of values (e.g., relative response
factors).

The analysis result cannot be
calculated because an operand value is
qualified (e.g., identifies analytes
whose internal standard is not found)
(Form 1 "X" Flag).

Analysis result is from a diluted
sample (DL suffix) and may be less
accurate than the result from an
undiluted sample (Form 1 "D" Flag).

Actual value is known to be greater
than the upper calibration range (Form
1 "E" Flag).

The analysis result is less than the
sample guantitation limit (Form 1 "J"
Flag).

The indicated analyte is a tentatively
identified analyte; its concentration
has been estimated (Form  1-E or 1-F
"J" Flag).

Results for the analyte are rejected by
the laboratory.

The indicated compound is an internal
standard.

A combination of "STD" and "BDL".
The indicated compound was found in the
associated method blank (LRB) as well
as the sample (Form 1 "B" Flag).
A Combination of "TIE1
1-E or 1-F "B" Flag).
and "FBK"  (Form
PRE  PRESUMPTIVE PRESENCE
Labels a suspected Aldol Condensation-
product for TICs  (Form 1-F "A" Flag).

Results of two or more injections are
not comparable (Form 1-D "P" flag),
e.g., Aroclor target analyte with
greater than 25% difference between
mean concentrations of the two column
analyses.

Presumptive evidence of presence of
material for TIC (Form 1-E or 1-F "N"
Flag).
                            H-35
                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
               APPENDIX A — FORMAT OF RECORDS FOR SPECIFIC USES
The USEPA does not warrant or  guarantee the completeness and/or accuracy of the
representative examples of  record  type  uses provided in this  appendix.   This
appendix  serves  as an example  for the  usage  of record types  and  in  no way
redefines or supersedes the specifications or requirements stated in Exhibits A
through H of OLM03.0.   NOTE:   Examples are representative and are not typically
provided for both columns used in the pesticide analysis.
                                     H-36                              OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
               Appendix A — Format of Records for Specific Uses

                               Table of Contents

Section                                                                   Page

1.0   VOLATILES	   40
      1.1   Record Type 10  for Volatiles	   40
      1.2   Record Type 11  for Volatiles	   40
      1.3   Record Type 20s and 30s for Volatiles —
            Initial Calibration Mean Values  (QC Code  'MNC')	   40
      1.4   Record Type 20s and 30s for Volatiles —
            BFB Tune  (QC Code 'LPC')   	   40
      1.5   Record Type 20s and 30s for Volatiles —
            Initial Calibration Standard (QC Code 'CLM')   	   40
      1.6   Record Type 20s and 30s for Volatiles —
            Dual Purpose Calibration Standard (QC Code  'CLD')  	   41
      1.7   Record Type 20s and 30s for Volatiles —
            Continuing Calibration Standard  (QC Code  'CLC') 	   41
      1.8   Record Type 20s and 30s for Volatiles —
            Method Blank (QC Code  'LRB')   	   42
      1.9   Record Type 20s and 30s for Volatiles —
            Storage Blank (QC Code 'LHB')	   43
      1.10  Record Type 20s and 30s for Volatiles —
            Instrument Blank (QC Code  'LIB')  	   44
      1.11  Record Type 20s and 30s for Volatiles —
            Regular Field Sample (QC Code  field Lt blank)  	   45
      1.12  Record Type 20s and 30s for Volatiles —
            Field Sample chosen for MS/MSD (QC Code 'LSD')  	   46
      1.13  Record Type 20s and 30s for Volatiles —
            Matrix Spike Sample (QC Code 'LF1')  	   46
      1.14  Record Type 20s and 30s for Volatiles —
            Matrix Spike Duplicate Sample  (QC Code 'LF2')  	   48

2.0   SEMIVOLATILES	   50
      2.1   Record Type 10 for Semivolatiles	   50
      2.2   Record Type 11 for Semivolatiles	   50
      2.3   Record Type 20s and 30s for Semivolatiles —
            Initial Calibration Mean Values (QC Code  'MNC') 	   50
      2.4   Record Type 20s and 30s for Semivolatiles —
            DFTPP Tune (QC Code 'LPC')  	   50
      2.5   Record Type 20s and 30s for Semivolatiles —
            Initial Calibration Standard (QC Code 'CLM')   	   50
      2.6   Record Type 20s and 30s for Semivolatiles —
            Dual Purpose Calibration Standard (QC Code  'CLD')  	   51
      2.7   Record Type 20s and 30s for Semivolatiles —
            Continuing Calibration Standard (QC Code  'CLC') 	   52
      2.8   Record Type 20s and 30s for Semivolatiles —
            Method Blank (QC Code  'LRB')   	   53
      2.9   Record Type 20s and 30s for Semivolatiles —
            Regular Field Sample (QC Code field is blank)  	   54
                                     H-37                              OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
               Appendix  A  —  Format of Records  for Specific Uses

                        Table of Contents  (continued)

Section                                                                   Page

      2.10  Record Type 20s and 30s for Semivolatiles —
            Field Sample Chosen for MS/MSD (QC Code 'LSD')	   55
      2.11  Record Type 20s and 30s for Semivolatiles —
            Matrix Spike Sample (QC Code 'LF1') 	   56
      2.12  Record Type 20s and 30s for Semivolatiles —
            Matrix Spike Duplicate Sample (QC Code 'LF2') 	   58

3.0   PESTICIDES (COL. 1)	   60
      3.1   Record Type 10 for Pesticides (Col 1)	   60
      3.2   Record Type 11 for Pesticides (Col 1)	   60
      3.3   Record Type 20s and 30s for Pesticides (Col 1) —
            Initial Calibration Mean Values (QC Code 'MNC')  	   60
      3.4   Record Type 20s and 30s for Pesticides (Col 1) —
            Florisil Cartridge Check Recovery Values (QC Code 'FLO')  .  .   61
      3.5   Record Type 20s and 30s for Pesticides (Col 1) —
            GPC Recovery Values (QC Code 'GPC')	   62
      3.6   Record Type 20s and 30s for Pesticides (Col 1) —
            Resolution Check Standard (QC Code 'LPC')  	   62
      3.7   Record Type 20s and 30s for Pesticides (Col 1) —
            Performance Evaluation  Mixture (QC Code 'CLE')  	   63
      3.8   Record Type 20s and 30s for Pesticides (Col 1) —
            Initial Calibration Multicomponent Standard (QC Code 'CLS')  .   64
      3.9   Record Type 20s and 30s for Pesticides (Col 1) —
            Initial Calibration Single  Component — Individual Standard A
            (QC Code 'CLM')	   65
      3.10  Record Type 20s and 30s for Pesticides (Col 1) —
            Initial Calibration Single  Component — Individual Standard
            Mix B (QC Code 'CLM')	   66
      3.11  Record Type 20s and 30s for Pesticides (Col 1) —
            Instrument Blank (QC Code 'LIB')   	   67
      3.12  Record Type 20s and 30s for Pesticides (Col 1) —
            Method Blank (QC Code 'LRB')  	   68
      3.13  Record Type 20s and 30s for Pesticides (Col 1) —
            Matrix Spike Sample (QC Code 'LF1') 	   68
      3.14  Record Type 20s and 30s for Pesticides (Col 1) —
            Matrix Spike Duplicate  Sample (QC Code 'LF2') 	   69
      3.15  Record Type 20s and 30s for Pesticides (Col 1) —
            Continuing Performance  Check — Individual Standard Mix A
            (QC Code 'CLE')	   71
      3.16  Record Type 20s and 30s for Pesticides (Col 1) —
            Continuing Performance  Check — Individual Standard Mix B
            (QC Code 'CLE')	   71
      3.17  Record Type 20s and 30s for Pesticides (Col 1) —
            Continuing Performance  Check — Performance Evaluation
            Mixture (QC Code 'CLE')	   72
      3.18  Record Type 20s and 30s for Pesticides (Col 1) —
            Field Sample chosen for MS/MSD (QC Code 'LSD')	   73


                                     H-38                              OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
               Appendix A — Format of Records  for Specific  Uses

                         Table of Contents (continued)

Section                                                                   Page

4.0   PESTICIDES (COL. 2)	   75
      4.1   Record Type 10 for Pesticides (Col 2)	   75
      4.2   Record Type 11 for Pesticides (Col 2)	   75
      4.3   Record Type 20s and 30s for Pesticides (Col 2) —
            Initial Calibration Mean Values (QC Code 'MNC')  	   75
      4.4   Record Type 20s and 30s for Pesticides (Col 2) —
            Resolution Check Standard  (QC Code 'LPC') 	   76
      4.5   Record Type 20s and 30s for Pesticides (Col 2) —
            Performance Evaluation Mixture (QC Code 'CLE')  	   77
      4.6   Record Type 20s and 30s for Pesticides (Col 2) —
            Initial Calibration Multicomponent Standard (QC Code 'CLS')  .   78
      4.7   Record Type 20s and 30s for Pesticides (Col 2) —
            Initial Calibration Single Component — Individual Standard A
            (QC Code  'CLM')	   78
      4.8   Record Type 20s and 30s for Pesticides (Col 2) —
            Instrument Blank  (QC Code  'LIB')   	   80
      4.9   Record Type 20s and 30s for Pesticides (Col 2) —
            Method Blank (QC Code  'LRB')  	   80
      4.10  Record Type 20s and 30s for Pesticides —
            Matrix Spike Sample (QC Code  'LF1') 	   81
      4.11  Rec rd Type 20s and 30s for Pesticides —
            Matrix Spike Duplicate Sample (QC Code 'LF2')  	   82
      4.12  Record Type 20s and 30s for Pesticides (Col 2) —
            Calibration Verification Multicomponent Standard
            (QC Code  'CLE' )	   84
      4.13  Record Type 20s and 30s for Pesticides (Col 2) —
            Field Sample chosen for MS/MSD (QC Code 'LSD')	   84

5.0   TYPE 90 — COMMENTS RECORD	   86
                                     H-39                              OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
 1.0   VOLATILES

 1.1   Record Type 10 for Volatiles

      10||||||GC/MS|OLMOI.OV||TESLPB||||esoooooi|INSTVI||TEST    LABS
      INC.|[000012C33

 1.2   Record Type 11 for Volatiles

      11JDB-5)|0.53|||||||||||000024CD9

 1.3   Record Type 20s and 30s  for Volatiles —  Initial  Calibration Mean Values
      (QC Code 'MNC')

      20|||o|MNC||18000|X1201|||||||||36|||000036668


      30|c|?4873| | |AVGJ0.53l| | | | | |R|l7.0| || M ! I I ! I ! I I I0000486BE


      30|c|?4839| | |AVG|l.536| | | | | |R|11.1 | | | | | | | | | | | |  |00005A717
1.4   Record Type 20s and 30s for  Volatiles — BFB Tune  {QC Code  'LPC')

      20|||0|LPC||18000|X1201|92|04|25|01|01||||l|||0004011E3


      2l|N||||||BFBOl|||||||||||0004121B2


      30Jc|460004||||I||  HI  |||j |||||| j||00042388C


      36|M| | |50 I |18.8|75|45.5| |95|100.0| | 96 | 7.2| |173|0.0|0.0|174|72.8| |175|5.7
      |7.8j176|71.6|98.3j177|4.9|6.8||||j  |||||||||000439122

1.5   Record Type 20s and 30s for Volatiles — Initial  Calibration Standard (QC
      Code  'CLM')

      20||VSTD20|0|CLM||18000|X1201|92|03|25|09|42||||36|||00044B8F2


      2l|N|||||(STDOI)  I)]  |||||00045C944


      23|P|92|04|25|01|011BFB011| | | | | | || ! I I I I I I M I I I I I I I I00046EA64


      30|c|74873|74975  | |0.563| |A|50.000| | | | | | || M I I ! ! I I I |000470F21




                                     H-40                              OLM03.0

-------
30|c|74839|74975| | |l.528| JA|50.000|
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
                                                          0004833E2
1.6   Record Type 20s and 30s for Volatiles — Dual Purpose Calibration standard
      (QC Code 'CLD')

      20||VSTD050|0|CLD| |18000|X1201|92|05|25|10|25| I I I 39 I I |0014300AA

      2l|N|||||(CSTDOI)|||||||||(00144117D

      23|p|92|05|30|05|oi|BFB02| | | |  | | | | | | | |  | | | ! I  I I I I I I !  |0014532A2

      30|c|74873|74975| | |o.93l| |A|50.000| | | |  || | ||  I ! I ! I I I  I I001465B66

      30|CJ74839|74975|j |1.821||A|50.000j ||||||||||||||]|J00147842C
      30|c|l7060070J74975|||2.089||A|50.000|
      30 C 3114554   STD   A 50.0
      32J|JRT|l5.74|
                        001833EF5
30|c|540363| | | STD | | JA|50.o|
32| | |RTJ9.97|
                             0018570D13
30|c|74975| | | STD | | |A|50.0|
32   RT 8. 00
                            |00187A10C
                                                       001810B47
                                          |l|l0928l|001822EDB
                                                  125103 00184618CB
                                                 26084 0018691EDC
1.7   Record Type 20s and 30s for Volatiles — Continuing Calibration Standard
      (QC Code 'CLC')

      20| |VSTD050|0|CLC| |18000|X1201 | 92|05 | 30|05 | 25 | | | |39| | |0024300AD
21 N
                CSTDOI
|00244117F
                                     H-41
                                                                       OLM03.0

-------
•Exhibit H — Appendix A
Format of Records for Specific Uses
      23|P|92|05|30|05|011BFB02| | I ! I I I II M I I I I I I I I I II I I I I0024532A1


      30|c|74873|74975| |  |0.96l| |A|50.000| | || I I I II I I II |F|81.0| | |002465B65


      30|c|74839|74975| |  |l.803| |A|50.000| | | | | I I I I I I I I |F|l7.4J | |00247842A
      30Jc|l7060070|74975| | J2.089J |A|50.000| | | | || I I I I M I |F|l0.6| | |002810B46


      30|G|3114554|  | |STD| | |Al50-°I M I I I I M I  I I I I I  |111092811002822EDA


      32)||RT|l5.74J||| M II M|002833EF3


      30|c|540363| |  |STD| | |AJ50.o| | |  || I M II I  II I I |I|125103|00284618C


      32|||RTJ9.97||||lI M M |0028570D1


      30|c|74975| | |STD| | |A|50.o| | | |  || 1 I II  II  II I III 26084|0028691ED


      32| |  |RT|8.00|  | |  | | | || M 00287A10D

1.8   Record Type 20s  and 30s for Volatiles  —  Method  Blank (QC Code 'LRB')

      20||VBLK01|1|LRB| 18000|X1201|92|05|30|06|11||ML|5.0|39|||00288CB9F


      2l|N|L| | | |  |FBLKOI| || I M I I  M |00289DCF6


      22|92|05|30|05|25|CSTD01|||||l.OJ|||00290F6D3


      23|p|92|05|30|05|oi|BFB02|MIMIMII!IIMIIPI0ITI°MII|002911B17

      30|CJ74873| |UG/L|BDL| |u| |  |u|lO| | || I  I I  I M M  I I I |002923881


      30|c|74839| |UG/L|BDL| |u| |  |u|lo| | || I  I M I I M  I II |0029355ED
      30|c|2037265|3114554|UG/L| |47.9| JA|50.0| || I II I M ISI96I II I I I |003251E4B

      30|c|460004|3114554|UG/L| |47.0| JA|50.0| | || M M I IsI94i  Mill |0032647E1
                                     H-42                              OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
30JCJ17060070|74975|UG/L| |50.3J |A|50.0J

30|c|3114554| | | STD | | |AJ50.o| | | | | j | | | | | j
                                                            | | | ] | J003277241

                                                   104439 0032895DF
32 | | j RT 1 15 . 70 |
                              00329A5EA
      30|c|540363| | JSTDJ | |AJ50.o| | | ! I! I I I I I I I I I 1 1 1 135938 | 00330C8A8
32 | | | RT | 9 . 92 | | )
                             00331D7DE
30|c|74975J | | STD j | |AJ50.0|
                                                 26488 00332F90C
      32J | JRTJ8.00| | | | j | | | j | (00333082C

1.9   Record Type 20s and 30s for Volatiles — Storage Blank (QC Code 'LHB')

      20 I IvHBLKOll 1 I LHB I I 18000 I X1201 I 92 I 05 I 30 I 06 I 21 1 I ML I 5.0 I 39 I I I 00688CB9F
      2i|N|L||||IFHBLKOI
                             00689DCF6
      22|92|05|30|05|25|cSTDOl|||||l.OJ |||00690F6D3
      23|P|92|05|30|05|01|BFB02 | | | | | |
      |006911B17


      30|c|74873||UG/L|BDL||U|||U|10|


      30|c|74839||UG/L|BDL||U| j|U|10|
                                   I MB I 92 I 05 I 30 I 06 I 111FBLK01|P|0|T|0|



                                   |||||||||||006923881

                                   IIIIIIMIII0069355ED
30 | C| 2037265 | 3114554 | UG/L | | 47 . 9 | | A | 50. 0 |


30 |c| 460004 | 3114554 | UG/LJ | 47 . 0 | | A| 50.0 | |


30 |C | 17060070 | 74975 | UG/L | | 50 . 3 | | A| 50.0 | |


30 | c| 3114554 | | | STD | | | A| 50 . 0 | | | | | |  | | | |  | | |
                                               | | S | 96 |

                                                S 94
                                                                | J007251E4B

                                                                 0072647E1
                                                     |s|lOl| | | | | | | 007277241
                                                 I 104439 0072895DF
      32RTis.70
                        00729A5EA
30|c|540363|||STD ||AJ50.0
                                                I13593800730C8A8
                                     H-43
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
32   RT 9. 92
                            (00731D7DE
30|CJ74975| | | STD j  | JA|50.0|
                                                 26488 00732F90C
      32|||RT|8.00||||||[j|||00733082C

1.10  Record Type 20s and 30s  for Volatiles — Instrument Blank (QC Code 'LIB')

      20||viBLKOl|l|LIB||l8000|xi20l|92|05J30|06|30||MLJ5.0|39|I|00688CB9F

      2l|N|LJ |||IFIBLKOI)||||||||||00689DCF6

      22|92|05|30|05|25|CSTD01|||||l.OJ |||00690F6D3
23 IP|92|05 I 30|05|011BFB02| | | | | | |
J006911B17


30|C|74873||UG/L|BDL||u|||u|lo||


30|C|74839||UG/L|BDL||U|||U|10| J
                                         |MB|92|05|30|06|ll|FBLKOl|p|o|T|o|
                                                   [006923881


                                                    0069355ED
30|c|2037265|3114554|uG/L| |47.9| |A|50.0|


30|c|460004|3114554|UG/L||47.o||A|50.0||


30|C|17060070|74975|UG/L||50.3||AJ50.0||


30|c|3114554|||STD|||AJ50.0|||||||||||||
32   RT 15. 70
                              00729A5EA
30 | C| 540363 | | | STD | | | A | 50 . 0 |
32| | |RT|9.92|
                             00731D7DE
30|c|74975| | |STO| | |A|50.0|
      32   RT 8.00
                       00733082C
                                                      |s|96| | | | | | |007251E4B

                                                      s|94|||||||0072647E1
                                                S 101
                                                                  007277241
                                                 I 104439 0072895DF
                                                1 135938 00730C8A8
                                                 26488 00732F90C
                                     H-44
                                                                 OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
1.11  Record Type 20s and  30s  for Volatiles — Regular Field  Sample  (QC Code
      field is blank)

      20||xi200|l|||l8000|xi20l|92|05|30|06|37J|ML|5.o|39|||0033432DB


      2l|N|L|||||FX1200|||[|92|05|2?|||J003354C2D


      22|92|05|30 JOB]25|CSTDOl| | | | |l.0| | | J00336660A


      23|P|92|05|30|05|01|BFB02|HB|92|05|30|06|21|FHBLK01||||MB|92|05|30|06|11
      |FBLK01JP|0|TJ2||||0|003379931
      30Jc|74873||UG/L|BDL||u|j|U|10|


      30|c|74839  |UG/L|BDL||u|[|U|10|
                                              |00338B69B


                                              I00339D407
30 JC | 2037265 | 3 114554 UG/L
                                 A50.0
30|c|460004|3114554|uG/LJ | 59 . 0 | |A|50.o|


30|c|l7060070|74975|uG/L| | 54 . 3 | | A| 50.0 |


30c 3114554   STD   A 50 . 0
      32    RT  15.66
                       J00275277B
30 C 540363   STD   A 50.0
32| | |RTJ9.9l|
                             002775948
      30 | C| 74975 | | | STD | | | A| 50. 0 |
32| | |RT|7.96|
                             00279898B
      30 | c| 124389 | | UG/L j TIE | 9 | JN |
32   RT 2 . 90
                             00281898B
     33 | CARBON DIOXIDE | 00281898B


     30 jc| 74630527 | | UG/L | TIE | 1 | JN |
                                                | S | 102 |


                                               | S | 98 |  | |


                                                S 107
                                                                | | JOQ271A07D


                                                                J00272CA19


                                                                  00273F48C
                                                 I  96179  002741762
                                                I  123502  002764A14
                                        [l|2633l|002787A4A
                                              002807A4A
                                                002827A4A
                                    H-45
                                                                       OLM03.0

-------
Exhibit  H  —  Appendix A
Format of  Records  for Specific Uses
 32   RT  25. 20
                              00283898B
       33|3-UNDECENE,  6-METHYL-,  (E) - | 00284898B

1.12   Record  Type  20s and 30s for Volatiles — Field Sample chosen  for MS/MSD
       (QC Code  'LSD' )

       20 | |X!20l|l|LSD| | 18000 | X1201 | 92 | 05 | 30 | 06 | 37 | |MLJ5.o|39| | J0033432DB
21 N L
           FX1201LSD
                         92 05 27
                                           003354C2D
22 92 05 30 05 25 CSTD01
                                   1 . 0    00336660A
      23 | P | 92 | 05 | 30 | 05 | 01 1 BFB02 | HB | 92 | 05 | 30 | 06 j 21 1 FHBLK01 1 | | | MB | 92 | 05 | 30 | 06 1 11
      | FBLK01 j P | 0 | T | 0 | | |  | 0 | 003379931
      30|C|74873| |UG/L jBDL| |U| |  |U110| j


      30|C|74839||UG/L|BDL||U|||U|10||
                                              [00338B69B

                                              I00339D407
30|c|2037265|3114554|UG/L| |50.8| JA|50.0|


30|CJ460004|3114554|UG/L| J49.0| JAJ50.0J  |


30|c|l7060070|74975|UG/L| |53.3| |A|50.0|  |


30 c 3114554   STD   A50.o
      32| j|RT|l5.65|
                        00375277B
30 | C| 540363 | | | STD | | | A| 50. 0 |
                                                       | S | 102 | | | | | | J00371A07D
                                                s98
                                                S 107
                                                                 00372CA19
                                                                  00373F48C
                                                   96178 003741762
                                                I 123501 003764A14
32J j JRT|9.90J
                            |003775948
30 | c| 74975 | | | STD | | |ft| 50. 0 |
                                               I 26330 003787A4A
32   RT 7. 95
                             00379898B
1.13  Record Type  20s  and 30s for Volatiles  — Matrix Spike  Sample (QC Code
      •LF1' )

      20| jX1201MS|l|LFl| | 18000 | X1201 | 92 | 05 | 30 | 07 | 14 | JML|5.0|39| | |00380B5C1
                                     H-46
                                                                 OLM03.0

-------
                                                 Exhibit H — Appendix A
                                     Format of Records for Specific Uses
2l|N|L| | | | |FX1201LFl| | j | j 92 | 05 | 27 | | | (00381CF38

22 | 92 | 05 | 30 | 05 | 25 | CSTD01 | | | | |l.OJ | | J00382E915

23|p|92|05|30|05|Ol|BFB02|HBJ92|05|30|06|21JFHBLKOl| | | JMB | 92 j 05 | 30 j 06 | 11
[FBLKOI|P|O| | |s|o|RJo|oo383icEC
30|c|74873| J UG/L | BDL |  | U | | | U | 10 |
                                              003843A56
30|CJ75354|74975|UG/L||40||A|50.00|||||Pj80|JD|4|||{




30|c|79016|540363|UG/L| |46| |A|50.00| | | | |P|92 | | D | 4| | |




30|G)71432|540363|UG/L||46||A|50.00|||||P|92||D|2|||




30|c|108883|3114554|UG/L||45||A|50.00|||||Pj91||D|6|


30|c|l08907|3114554|UG/L|  |44 | |AJ50.00| | |  | |P|88| |o|7|
                                                         003911212
                                                          004026477
                                                          00405C84B
                                                            00412ABC4
                                                            00413D5FF
30 | C| 2037265 | 3114554 JUG/L | ] 46 . 1 | | A| 50.0 |
                                                | S | 92 | | | | j | |004175BA7
30|c|460004|3114554|UG/L|
A 50 . 0
                                                S 93
                                                           004188542
30 |C | 17060070 | 749 75 | UG/L | | 54 . 3 | | A| 50. 0 |


30 |c| 3114554 | | ] STD | | | A | 50. 0 | | |  | | |  | | |  | | |
                                               | S | 109 | | | | j | j 00419AFC2
                                           I 101076 00420D350
32   RT 15. 70
                        00421E35B
30|c|540363J | | STD | | |A|50.0|
                                            125573 004220613
32   RT 9. 92
                       00423 1549
30|c|74975| | | STD | | | A | 50. 0 |
                                           26701 004243657
                               H-47
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
      32) | |RT|7.98J | | || I I I I I J00425459B

1.14  Record Type 20s and 30s for Volatiles — Matrix Spike Duplicate Sample (QC
      Code ' LF2 ' )

      20 | |xi201MSD|l|LF2J | 18000 | X1201 | 92 | 05 | 30 | 07 | 55 | |ML|5.0J39| | J00426726B

      2l|NJL| | | | |FX1201LF2| | | | |92|05|27| | | |004278BE3

      22 | 92 | 05 | 30 05 | 25 | CSTD01 | | | | |l.OJ | | J00428A5CO


      23 | P | 92 | 05 | 30 | 05 | 01 j BFB02 | HB | 92 | 05 | 30 | 06 | 21 1 FHBLK01 1 | | | MB | 92 | 05 | 30 | 06 1 11
       FBLK01 P 0   S  0 R  0  00429D997
30|c|74873| | UG/L | BDL | | U |  | | U | 10 |
                                                    00430F701
      30|CI 75354 I 74975 IUG/LI |38 |A|50.00J | | | JP| |77|D|4J | | |




      30|c|79016|540363|uG/L||44|  A  50.00  ||||P||88|D|4|||




      30|C|71432|540363|UG/L||47|  A  50.00||||  p||94|DJ2|||




      30|c|l08883|3114554|uG/L|[48||A|50.00|||||P||97|o|6|


      30|c|l08907|3114554|uG/L||47||A|50.00|||||P||94|o|7|
                                                        (00437CED4




                                                        ||004482146




                                                        ||004518527




                                                        ||||0045868B3


                                                        I|II0045992EE
30 | C| 2037265 | 3114554 |UG/L | | 49 . 2 | | A | 50. 0 |


30|c|460004|3114554|UG/L| |47.6| |AJ50.0| |


30 | C| 17060070 | 7497 5 JUG/LJ |55.2||A|50.0||


30 | C| 3114554 | | | STD | | | A | 50 . 0 | | |  | | | | | | |  | | |


32| |  JRT|l5.70| | |  | | |  Ml |00467A055
                                                | | | | | | | S | 98 | | | | | | |0046318AA


                                                | | | | | | S | 95 | | | | | | |004644251


                                                ||||||s|llO|||||| |O04656CBF


                                                 I  101145  00466904A
                               H-48
                                                                       OLM03.0

-------
                                                 Exhibit H — Appendix A
                                     Format of Records for Specific Uses
30JC|540363|||STD|||A|50.o|||||||J||||||||I|124184|00468C30A


32|||RT|9.94|||  |||||||00469D24C


30|CJ74975|  | |STD| | |A|50.o| | | | || | | || I I i I I 11126683|00470F36D

32|||RTJ7.98|||||||||||0047102B1
                               H-49                              OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
2 . 0   SEMIVOLATILES

2.1   Record Type 10 for Semivolatiles

      io| | | | |  IGC/MS | OLMOI.OB | |TESLAB| | | | esoooooi | INSTBI | (TEST    LABS
      INC. | |000002BA9

2.2   Record Type 11 for Semivolatiles

      11 | XRT-1 | | 0 . 52 | | | | | | | | | | | 000024C8D

2.3   Record Type  20s  and  30s for  Semivolatiles  — Initial Calibration Mean
      Values (QC Code 'MNC')
      20 | | |o|MNC| J18000|X1201J
                                      72    00003661C
      30|c|l08952| | |AVG|l.817| | | | | (RJ7.2


      30 | c| 111444 | | |AVG|l.607| |||||R|6.2|


      3G|c|95578| | |AVG|l.375| | | | | JR|3.l| |


      30|c|54173l| | |AVG|l.502| | | | | JR|3.l|
                                                     | 000048677


                                                     J00005A6C4


                                                     00006C650


                                                      00007E685
2.5
      30 | c| 93951736 | | |AVG|l.22l|
                                                       000745BDE
      30|c|219969l| | |AVG|o.91l| | | | | |R|2.o| | | || I ! I I I I I I |000757DOA

2.4   Record Type 20s and 30s for  Semivolatiles  —  DFTPP Tune  (QC Code  'LPC' )

      20 | | |O|LPC| I 18000 I X1201 1 92 I 05 I 01 1 00 I 32 I | | |l| | | 000769E6E
      21 | | | | | | |DFTPPOI|
      30 C 5074715
                           | | | | | |00077AF43


                           I I I I I I I I I I I I I I I I 00078C71C
      36JM| j | 51 | | 41. 9 | 68 | 0.5 | 1.3 | 69 | 37. 6 | | 70 | 0.0 | 0. 0 | 127 | 45 . 2 | |l97|0.4J |l98|lO
      0.0 | 1 199 | 8.0 | | 275 | 25.0 | J365|3.69| J44l|l3.7J |442|99.l| | 443 | | 20. 8 j 21 .0 1 000
      793554

      Record  Type  20s and 30s for Semivolatiles —  Initial  Calibration Standard
      (QC Code 'CLM' )

      20 | |SSTD080JO|CLM| | 18000 | X1201 | 92 | 05 | 01 | 02 | 01 | | | ]72| | |00155B94D
                                     H-50
                                                                      OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
21
          STD02
                               00156C8F3
      23 | P | 92 j 05 | 01 j 00 | 32 | DFTPP01 | | I I I I 1 I I I I I I I I I 1 I I 1 I I I I I I 00157EBBA

      30 |c| 108952 | 3855821 1 |  | 1.853 | |A|40.000| | | | | | | | | || II II I !  |0015812F5


      30|cj 111444 | 3855821 1 |  | 1.599) \h\ 40.000 | || II I I I M I I I I I M  |001593A37
2.6
30|c|93951736|385582l|||l.248J|A|40.000J|||


30|c|219969l|3855821  ||o.928||A|40.000|| j ||

Record  Type  20s and  30s  for Semivolatiles
Standard  (QC Code  'CLD')
                                                              |00228C51D
                                               | | | I I I | | |00229ED4C

                                               Dual Purpose  Calibration
      20  SSTD50 0 CLD  18000 X1201 92 05 01 02 40    72   0045914F6
      21|!J||||STD03||
                         00460249D
      23P9205010032DFTPP01
      30JCJ108952|385582l|  |l.462||A 25.000|

      30|C|111444|3855821|||1.233[|A|25.000|
                                                     [004614764


                                                     |||004623B3A


                                                       00463664C
      30|c|219969l|3855821 ||0.828||A|25.000|


      30|c|ll46652|||STDj|JA|20.0|||||||||||I
      32)||RT|7.23|
                       005354AE6
30|CJ15067262|||STD|||A|20.0||)|


32) | |RT|9.17J  | | I I I I I I I |005377DA2


30|c)l517222|||STD||]A|20.0||||[j
      32   RT10.82
                       (00539B042

                               H-51
                                            |||||||||||||0053318FO

                                           I|58474|005343BC2
                                                  I27242005366E6F
                                                |l|38472|00538A043
                                                                       OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
      30|c|l520963| | | STD | | |A|20.o| | | | | || I I I I I I I I 1 1 1 18770 | 00540D2F2
2.7
      32) | |RTJ17.95|
                              00541E310
      30 | CJ 1719035 | | | STD | | | A| 20. 0 |
      32   RT 14 . 22
      30 | c| 3855821 | | | STD |
      32RT5.87
                              0054315AF
                             00545479F
                                                 I 24292 0054205B2
                                                   17522 00544385F
      Record Type  20s and  30s  for  Semivolatiles
      Standard  (QC Code 'CLC')
—  Continuing  Calibration
      20 j |SSTD050|0|CLC| |18000|X1201192|06|01111130| | | |78| | |00459DC7F
      21 I I I  IIICCSTD01||  | | | |  III00460ED24
      23P9206  011101DFTPP02
      30 c 108952 3855821   l.462  A25.000
      30|c|lll444|385582l|||l.233||A|25.000|
                                                          |004610FEB
                                                         F 19 . 5    004623B3A
                                                         F23.300463664C
      30C219969138558210.828   A25.000
      30C1146652STD
      32RT7.23
                            |005354AE6
      32|||RT|9.17|
                             005377DA2
      30|c|l517222J | j STD | | JA|20.o|
      32RT10.82
                              00539B042
      32|||RT|l7.95|
                             |00541E310

                                     H-52
                                                  ||||||||F|9.1|||0053318FO


                                                  J58474|005343BC2
      30|c|l5067262| | |STD| | |A|20.o| | | | || I I 1 I I I I ! I 11127242|005366E6F
                                                   38472 00538A043
      30|c|l520963| | |STD | | |A|20.o| || ! I I I I I I I I I I I 11|18770|00540D2F2
                                                                       OLM03.0

-------
30 |C | 1719035 | | |STD| | | A| 20.0 |
32) |  |RT|l4.22|
                              |0054315AF
30|CJ 3855821 | | | STD | | | A | 20. 0 |
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
                                                 l242920054205B2
                                                 I 17522 00544385F
      32| | |RTJ5.87| | | |  | |  | | | | |00545479F

2.8   Record Type 20s  and 30s  for Semivolatiles  — Method Blank (QC Code 'LRB')

      20 | |SBLK32|l|LRBJ | 18000 | X1201 | 92 | 06 | 01 | 12 | 10 | JML | 1000 | 78 | ) | 005467308
21
                |FSBLKOl|92|03|30| | | | | |2.0| J005478CBB
      22|92|06|Ol|ll|30|cCSTDOl| |  1000 | 1 . 0 | j | | 00548AA6D
      23 P 92 06 01 11 01 DFTPP02
30 C 108952  UG/L BDL
                           U 10
                                             P 0 1 0
                                                         00549D056
                                                     00550EE8E
      30|c|4165600|l520963|UG/L| J40.9| |A|50.0|


      30|c|321608|l5067262|UG/LJ|42.8||A|50.0|


      30|c|98904439J1719035|UG/L||52.o||A|50.0


      30|c|4165622|385582l|UG/L| 67.2||A|75.0|


      30|C|36712" | 3855821 | UG/L |  |61.7J  |A|75.o|  |


      30|c|ll8796|l5067262|UG/L| 70.9||A|75.0|
                                           I I | I I |s|82| | | | | | (006147181


                                           | | | | I IsI86I I I II I (006159BF9

                                           |||||||s|l04|||||||00616C83A


                                           ||||||S|90|||||||00617F2D2


                                           |||||S|82|||||||006181C81


                                           ||||||s|95|||||||00619472C
30|c|93951736|3855821|UG/L||69.2||AJ75.0|j||||j  ||s|92||||||(0062072BF


30|c|219969l|385582l|UG/LJ|38.9||A|50.0||||||
                                                |s|78|
                                                                  006219D66
30C 1146652   STD   A 20 . 0
      32||(FT|7.72)|||||||||J00623CF6A
30c 15067262   STD   A 20. 0
                                                 1 94564 00622C038
                                                  1 46152 00624F2FE
                                     H-53
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
2.9
32   RT 9. 65
                             00625023E
30 |c| 1517222 | | | STD | | | A | 20. 0 |
                                                 I 65936 0062624EE
32) | |RT|ll.32|
                             |0062734DF
      30|c|l520963| | | STD | | |A|20.0J || | | | | || I I I I I I 1 1 1 35768 | 00628579F
32 | | JRTJ 18.44)
                              0062967B8
      30Jc|l719035| | | STD | | |A|20.o| || I I I I I I I M I I I 1 1 1 43708 | 006308A67


      32) | |RT|l4.70| M M I I I I I |006319A71


      30|c| 3855821 | | | STD | | | A I 20 • ° I I M I I I I I I I I I I I 1 1 1 29752 | 00632BD33
      32|||RT|6.35
                       00633CC63
Record Type 20s and 30s for Semivolatiles — Regular Field Sample (QC Code
field is blank)
20
21
                   |l8000|xi20l|92|06|0l|l2|20| |ML|lOOO|78| | J00434F7E4
       c   FX120l920530
                                            2.0  00435 18E4
      22|92|06|Ol|ll|30|cCSTDOl| | | |lOOO|l.O| | | |004363696
      23 [ P | 92 | 06 | 01 | 11 | 01 | DFTPP02 |
      | | (OJ004376C26
      30 1 cl 108952 1 |UG/L|BDL| lul I luliol
                                     I MB[92 I 04 I 011 15 114 IFSBLK011P|0|T|0|
                                               004388A5E
30|cJ4165600|l520963|UG/L||43.0||A|50.0||


30|c|321608|l5067262|uG/L||48.3||A|50.o||


30|c|98904439|l719035|UG/L||51.8||AJ50.0|


30|c|4165622|385582l|UG/L||69.8||A|75.0||


30|c|367124|385582l|UG/L| |69.6| |A|75.o| | |



                               H-54
                                                       S 86
                                                 s97
                                                            005020B12
                                                                  00503358D
                                                 | | | | | | |s|l04| | | | | | |0050461DF


                                                 | I I I IIs I 93I I I I I I |005058C8C
                                                      s|93| | | | | | |00506B64E
                                                                       OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
30
30
30
30
32
30
32
30
32
30
32
30
32
30
32
c
c
c
c

c

c

c

c

c

118796J 15067262 |UG/L| 79.5 A 75.0 S 106 | 00507E1D3
93951736
2199691
3855821
RT
6.38
1146652
RT
7.75
385582l|UG/L| 73.0 A 75.0 |sJ97 | 005080D64
3855821 |UG/L| 45.6 A 50.0 S 91 | 005093801




15067262
RT
9.67
1517222
RT


11.34
1719035
RT

14.74
1520963
RT

18.49












STD

STD

STD
i
STDJ

STD

STD

A 20.0 I 30288 005105AB5
0051169E8
A 20.0 I 96100 005128C90
005139BCF
A 20.0 | I 47432 |00514BF65
00515CEA7
A 20.0 || |l 71076 00516F145
005170142
A 20.0 I 36728 0051823F5
00519340D
A 20.0 | I 27356 0052056BD
0052166DB
2.10  Record Type  20s and  30s  for Semivolatiles  — Field  Sample  Chosen for
      MS/MSD (QC Code 'LSD')

      20||xi20l|l|LSDJ|l8000|xi20l|92J06|Ol|l2|50||ML|lOOO|78|||00634F7E4


      2l||L||c|||FX1201LSD|92|05|30||92|05|27||2.0JJ0063518E4


      22|92|06|Ol|ll|30|cCSTDOl||||lOOO|l.o||||006363696


      23|p|92|06|0l|ll|oi|DFTPP02|||||||||||MB|92|04|01|15|14|FSBLK01|P|0|T|0|
      ||JOJ006376C26


      30|C1108952| |UG/L|BDL j |U|  | |U110| | | | | |  | |  | | j   | | |  |006388A5E
                                     H-55                              OLM03.0

-------
 Exhibit H — Appendix A
 Format of Records for Specific Uses
30|c|4165600|l520963|UG/L| |43.0J |AJ50.0| | |  |


30|c|321608J15067262|UG/L||48.3||A|50.0||||


30JC|98904439|1719035|UG/L||si.8|JA|50.0|||


30|c|4165622J385582l|UG/L||69.8J|AJ75.0||||


30|c|367124|385582l|uG/L||69.6||AJ75.o|||||


30|C|118796|15067262|UG/L| | 79 . 5 | | A|75.0| | |  |


30|c|93951736|385582l|uG/L||73.0[|A|75.0|||


30|c|219969l|385582l|UG/L| |45.6| |A|50.0| | |  |


30|c|385582l| | |STD| | |A|20.o| | || I I I I M I I  I
                                                        S  86
                                                                   007020B12
                                                    I I II s I 97 I I I I I I I 00703358D


                                                    i I I I ISU°4| I I I I I |0070461DF

                                                    M I I S | 93 | | | | | | | 007058C8C


                                                    I I ISI93I I I I I I |00706B64E
                                                 S 106
                                                        S 97
                                                                    00707E1D3
                                                                    007080D64
                                                                   007093801
                                                 I 30288 007105AB5
32| | |RTJ6.38|
                             0071169E8
      30|c|ll46652| | | STD | | |A|20.o| | || I II I I I II I M I r I 96100 | 007128C90
32   RT 7. 75
                             |007139BCF
      30cl5067262   STD   A 20. 0
                                                    47432 00714BF65
32   RT 9. 67
                             |00715CEA7
      30|c|l517222| | | STD |  | |A|20.0J | | || 1 I I I I I I I I I | 1 1 71076 | 00716F145
      32   RT 11. 34
                              007170142
      30|c|l719035| | |STD| |A|20.0| | || I I ! I M I M M 11136728|0071823F5

      32||JRTJ14.74J ||||| ||| 00719340D


      30|c|l520963| | |STD| | |A|20.0J | || I I I I I  I I !  I I  I |1127356|0072056BD
32) | |RT|l8.49|
                              0072166DB
2.11  Record Type 20s and 30s  for  Semivolatiles — Matrix Spike Sample  (QC Code
      'LF1' )

      20| |xi201Ms|l|LFl|  | 18000 | X1201 1 92 | 06 | 01 1 13 | 34 | |ML|500|78| | |00722931C

                                     H-56                              OLM03.0

-------
                                                 Exhibit H — Appendix A
                                     Format of Records for Specific Uses
2l| |L| |c| | |FX1201LFl|92J05|30| |92|05|27| |2.C| J00723B441


22|92|06|oi|ll|30|cCSTDOl||||500|l.o||||00724D125
23|P|92 j 06|011111011DFTPP02|
0|R|OJ00725074C
                                      JMB|92|06|Ol|l2|lo|FSBLKOl|p|o|
30|c|108952|3855821|UG/L||61||A|75.00|||||P|82||D|2|||||||||007263177


30|C|111444||UG/L|BDL||U|||U|10||||||||||||||||007274FA5


30|c|95578|3855821|UG/L||61||A|75.00|||||P|81||D|1|||||||||00728790F
30 |c| 54 1731 |  | UG/L | BDL |
                             10
007299743
30|G|106467|3855821|UG/L||37||A|50.00|||||P|74 |o|l
                                                            |00730C173
30jJ|4165600|1520963|UG/L| |37.8| |A|50.0| | | ' | | | | | |3 | 76 | | | | | | |0079035AC


30|c|321608|l5067262|UG/L| |40.3| |A|50.o| j | | | | j | | |S | 81 | | | | | | |007916004







30|c|219969l|385582l|UG/L||38.4||A|50.o||||||||||s|77|||||||007976162


30|C|1146652|||STD|||A|20.0||||||||||||||||I|101044|0079884EO


32|||RTJ7.70J ||I|||||||007999410
30 |c| 1719035 | | | STD | | | A | 20 . 0 |
                                           I 37488 008064F42
32   RT 14.69
                        008075F5E
30 | CJ 3855821 | | | STD | | | A | 20. 0 |
                                           I 30512 0080881FE
                               H-57
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
32 | | |RT|6.33J
                             008099122
2.12  Record Type 20s and 30s for Semivolatiles  — Matrix Spike Duplicate Sample
      (QC Code 'LF2' )

      20 | | X1201MSD | 1 | LF2 | | 18000 | X1201 | 92 | 06 | 01 | 14 | 05 | | ML | 500 | 78 | | | 00810BDF7


      2l| |L| |c| | |FX1201LF2|92|05|30| |92|05|27| |2.o| JOOSIIDFID


      22 | 92 | 06 01 | 11 | 30 CCSTD01    500 1 . 0   00812FC01
23|p|92|06|oi|ll|oi|DFTPP02|||||||
0|R|0|008133228


30|c|l08952|385582l|UG/L||63|JA|75.00|


30|C|111444||UG/L|BDL||U|||U 10|||||||
                                           |MB|92|06|oi|l2|lo|FSBLKOl|p|o|
      30|c|541731||UG/L|BDL|
30|c|l06467|385582l|UG/L| |3?| |A|50.00|
                                                     oosi57A8F
      30JCI 95578 I 38558211UG/LI |6l||A|75.00|||j|p||82|D}l|
                                               00817C22E
                                                  |73|o|l|
                                                                  008145C61
                                                           00816A3FA
            00818EC53
      30|c|4165600|l520963|UG/L| |39.2J |A|50.0|
      30 | C| 32 1608 | 15067262 |UG/L| j 40. 1 | | A | 50. 0 |
 s78
Js|80|
                                                            0087865AD
                                                             008799002
      30|c|ll46652| | | STD | | |A|20.o|
32   RT 7 . 73
                             00887C42D
      30|CJ15067262J | | STD | | |AJ20.0|
32| | |RT|9.65|
                             00889F712
      30|c|l517222| | | STD | | |A|20.o|
                                      |||||I|100060|00886B4FA
                                           |l|50488J00888E7D2
                                           I 77500 0089019AE
                                    H-58
                                                                 OLM03.0

-------
                                                 Exhibit H — Appendix A
                                     Format of Records for Specific Uses
32 | | |RT|ll.32|
                        00891299F
30cl520963   STD
                                             29384 008924C52
32J | |RT|l8.44|
                        008935C6B
30|c|l719035| | | STD | | |A|20.0|
                                             39388 008947F23
32   RT 14. 70
                        008958F2D
30|c|385582l| | | STD | | |A|20.0| | | | 1 | | || I I I I ! I | 1 1 29976 | 00896B201
32| | |RT|6.37|
                      |00897C133
                               H-59
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
3.0   PESTICIDES (COL. 1)

3.1   Record Type 10 for Pesticides (Col 1)

      10!1111|GC|OLMOI.OP||TESLAB|11|68DOOOOi|iNSTPi|[TESTLABS INC.||ooooo2A30

3.2   Record Type 11 for Pesticides (Col 1)

      ll|DB-5||0.53||||||||||J0000249FA

3.3   Record Type 20s  and 30s for Pesticides (Col 1) — Initial Calibration Mean
      Values (QC Code 'MNC')

      20|||0|MNC||l8000JX120l|||||||||6l|||000036387
      30 C 319846   AVG 834490      R 6
                                                     000048346
      32RT6.92RTF6.87RTT6.97|000059DCB
30 C 319857   AVG 272332      R 12
      32|||RT|8.24|RTF|8.19|RTT|8.29 ||||| 00007D89C
                                                      00006BE34
30|c|5103742| | |AVG|706395| | | | | | R I 13 I
      32   RT 11. 87 | RTF | 11. 80 | RTT | 11.94|
      30|c|8001352 |||13757
      32   RT 14. 84 RTF | 14.77 | RTT | 14.9l|
      30c8001352   | 7373 |
      32   RT 15. 47 RTF 15 . 40 | RTT | 15 . 54
      30 C 8001352  | 1 10643 |
                                                       00042FD21
                                      ||000431AOA


                                      ||||00044363C


                                      |l|000455429


                                      | | |000466F72


                                      |2|000478D57


                                      I I I I00048A976
      32 | | | RT | 16 . 04 | RTF | 15 . 97 | RTT | 16 . 11 | | | | | | 3 | 00049C748


      30|c|8001352| | | |l7393| | || | | || | I I  I I I I ! I I II |00050E370


      32| | |RT|l7.23|RTF|l7.16|RTT|l7.30| | | | | |4|000510136
                                     H-60
                                                                        OLM03.0

-------
3.4
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
      30|C|11096825||||48393|||||||||
32   RTl4.0lRTF13.94RTTl4.08
30|c|877098| | |AVGJ344348| |  | | | |R | 10 |
                                           I00120C450
                                            J5J00121E21E
                                                      0012202BA
32 |  | | RT | 5. 04 | RTF | 4. 99 | RTT j 5. 09 | | j | j | (001231D26

30|c|2051243| | JAVGJ455516| |  j | |  |R|lO| || M M N I ! I I (001243E74

32 j  | | RT | 22. 53 | RTF | 22. 43 | RTT | 22. 63 | | j |  | |  |001255B43

Record Type 20s  and 30s for Pesticides (Col 1) — Florisil Cartridge Check
Recovery Values (QC Code 'FLO')
20| |FLOP1248Al|o|FLo| 18000 | X1201 1 91 1 11 1 14 | 09 [ 19 |
                                                              (00126852C
21
30
30
30
30
30
30
30
30
30
30
30

c
C
c
c
c
c
c
c
c
1


P1248A1
319846
58899
76448
959988
60571
72208
72548
50293
72435
c|877098
C
205124.
NG
NG
NG


NG
NG
NG
NG
NG
NG





NG



|8.695
8.816
8.690


|s. 891
17.498
16.576
23.206
16.551
84.668
9.882
3 |NG 20.
2.0
A 10
00127993B
.000
A 10.000J
A 10.000)
A 10
A 20
A 20
|A 20
A|20
000
000
000
000
000








A 100.000
A 10.
032 A :
000







1
1





10.000



1




1




















































S
S
S












87. 0|
88
86
S
S
S
<
5
S

.2
.9


88.9
87.5
82.9

L:
L6.
(





)
82. 8|
S
84.

















7 | |
S|98.8


S



100.2
1




1

1


1


00128BFC9
00129E581
001300B3E
|0013131F9
| 001325882
001337EF2
|00134A61A
00135CC78
00136F3EA
(001371A93
|

0013843
                                     H-61
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
3.5   Record Type 20s and 30s  for Pesticides  (Col 1) — GPC Recovery Values (QC
      Code ' GPC ' )
3.6
      20 | |GPC1242A1|OJGPCJ 1 18000 | X1201 1 91 1 11 1 13 | 09 | 19 |
21 | | | | | | |G1242Al|
                                . 0| |00140993B
30 | c| 58899 | | NG | |9.816J |A|lO.OOO| |


30|c|76448| | NG | |9.690| |AJ10.000| |


30|CJ309002| |NG| J9.89l| |A|lO.OOO|


30 | C| 60571 1 |NG | | 18. 498] |A|20.000|


30|c|?2208| |NG| |l7.576| |A|20.000|


30 |C | 50293 | |NG | | 17. 581 1 |A|20.000|
                                               S 98. 2
                                               s96.9
                                               |s|98.9|
                                                S 92 . 5
                                                s87.9
                                                S 87 . 9
                                                              00139852C




                                                           J00141E581


                                                           |001420B3E


                                                           | J0014331F9


                                                           | | 001445882


                                                           | |001457EF2


                                                             00146CC78
      Record Type  20s  and  30s  for  Pesticides  (Col  1)  — Resolution  Check
      Standard (QC Code 'LPC')

      201 |RESCOl|o|LPC| |l8000|xi20l|92 |04|20 04|44| | | |5 | | |001396901
21
               RESCOl
                              2.0  001407BA9
30|c|877098| | | |


32|||RT|5.06|||


30|c|5103742|||


32|||RT|ll.88||


30|c|959988 |||


32j ||RT|12.40||


30|C|72559|||J|


32|||RT|13.10||


30|c|6057l|llll


32|||RT|l3.26||
                          ]R|98.9| || ! I 1 I I ! I I I I |0014196FF
                             00142A62C
                           |R|IOO.O|
                         llll |00144D2EE


                         | |R|IOO.O| | | | |


                         | | | | |00146FEF7


                         |R|92.3|||||||


                         llll |00148292B


                         |R|IOO.O|


                              001505422
                                                 00143C2DE
                                                00145EEFC
                                           |||00147193C
                                               001494422
                                     H-62
                                                                       OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
30|c|l031078||||


32|||RT|l6.03|||


30|c|72435||||||


32|||RT|17.87|||


30|c|53494705|||


32)||RT|18.15|||||||
                          | | | i J0015280C6


                          |K|98. 6)1111111

                              00154AB36
                                                 0015170C8
                                        001539B17
                                                  |00155C8F1
                              00156D8FE
                                           00157FOAD
32   RT 22. 53
                              0015800AD
3.7   Record Type 20s and 30s for Pesticides (Col 1) — Performance Evaluation
      Mixture (QC Code 'CLE')

      20| JPEM01|0|CLE| |18000|xi20l|?2|04|20|06|26| I  I I12| I  I 001592687
      21||I|I|IPEMOI
        I I I I I I I
                      |2.0||00160389F
30 |c| 319846 | JNGJ J0.020| |AJ0.020|
      32RT6.94
                       001626C20
30|c|319857| | NG | |o.022| |A|o.020| | |R|94.0|
32 | j | RT | 8. 25 |
                             00164AOED
30|c|58899| | NG |  0.020| |ft|0.020| | |R|93.4| | | |DJO.O|
32| | JRT|8.03| | |


30|c|72208| |NG|
                             00166D388
                      |A|O.IOO| | |R|98.5|
                                                                001615CE1
                                                                  0016391BC
                                                               00165C465
                                                                 0017048D9
                                                          0017284BB
      32| | |RT|l4.26| | | | | | PB | 11.0 | 13. 5 | | |0017160EC


      30JCJ50293| |NG | | 0.201 | | A | 0.200 | | | R| 94 . 6 | | | | D | -0 . 5 |


      32| | |RT|l5.44| | | |  | PB | 2 . 5 | 13 . 5 | | |001739COC
      30|c|72435| | NG | |0.54l| JA|0.500| | |R|97.4| |  | |o|-8.2| || | M I I |00174C014
      32RT17.86
                        00175D032

                               H-63
                                                                       OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
      30|CJ877098| | NG | |o.038| |A|0.040| | |RJ92.6| | | | D | 5 . 0 | | | | | | | | J001806222
32 | | |RT|5.06|
                             00181714F
30|c|2051243| | NG | |0.042| |A|o.040| | j | | | | |o|-5.0| | | | | | | | |
                                                             0018296ED
32   RT 22 . 53
                              00183A6ED
3.8   Record Type  20s  and 30s for  Pesticides  (Col 1) —  Initial  Calibration
      Multicomponent Standard (QC Code 'CLS')

      20) |AR166001|0|CLSJ | 18000 | X1201 | 92 | 04 | 20 | 06 | 58 | | | | 12 | | |00184CFD8
21 | | | | | | |AR16600l|
                               | 2.0 | |00185E4C3
      30|c|l2674112| | | |27928| |A|o.200| | | || I I ! I ! I I I I I M |0018606AO
32
30
32
30
32
30
32
30
32
30
32
30
32
30
32
RT
9.00| |
C 12674112
RT
10.80
C 12674112
RT
11.14
C 12674112
RT
9.46
C 12674112
| RT
7.99
C 11096825
| RT|l6.86
C 11096825
RT
14.23
C 11096825
RT
18.14
1 00187169E
6941 A 0.200

2 001894872
6398 A 0.200

3 001917A4B
9007 A 0.200

4 00193AB5F
10821 A 0.200|

5 00195DD48
94264 A 0.200

1 001971038
74552 A 0.200

| 2 | 001994308
32609 A|0.200

3 0020175DO
                                                      001883797
                                                      001906971
                                                      001929B36
                                                      |00194CD18
                                                      00196FF3E
                                                      00198322C
                                                      0020064E5
                                    H-64
                                                                      OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
      30 | CJ 11096825 | | | | 34595 | |A| 0.200 | | | | || ! I I I I I I 1 I I I |0020297C7

      32 j |  | RT j 19 . 66 J | | | j j | | | | 4 | 00203A8CF


      30|c|ll096825| | | |48393J JA|0.200| | 1 || I I I I I I I I I ! I I |00204CABD
32   RT 14 . 01
                             5 00205DBA2
      30|c|877098| | | | 344348 | |AJ0.040| | | I I I I I I I I I I  I I  I ! |00206FCDA
32   RT 5. 05
                             002070C06
30|CJ2051243J | | |455516| JA|0.040J
                                                       002082DFO
      32RT22.54
                        002093DFB
3.9   Record Type  20s  and 30s  for  Pesticides  (Col 1)  —  Initial Calibration
      Single Component — Individual Standard A (QC Code 'CLM')

      20INDAL130CLM18000X1201920420114111   00304C753
      21|||||||lNDAL13|
                         2.0  00305DA70
      22
                1.0 00306E736
      30JC|319846| | | |867823| |AJ0.010|
32| | |RT|6.9l|
                             003081789
30|c|58899| | j |807087J |AJ0.010| | | | | I I I I I


32|||RTJ8.0l|||||||||||0031046FD


30|C|76448||||827300||A|0.010||||||||||
                                                003070857
                                                     0030937DC
                                                     003116728
      32RT8.89
                       00312766D
30|c|959988| | | |605148| |A|0.010|
32 | j | RT | 12 . 38 |
                              00314A7AB
                                                      0031397A9
30c6057l    638273  A 0.020
                                                     00315C7CC
      32)|(RTJ13.24
                        00316D7CA
      30JC|72208|j|J444069)|A|o.020||)|||||||||  |||||00317F7F3
                                     H-65
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
32 | | |RTJ14.25|
                              0031807FD
30|c|72548| | | |444235| |A|0.020| | |  || |


32| | |RTJ14.62| | | I I I I I I I |00320383D


30|c|50293| | | |401902| JA|0.020| ||  M I
                                                     003192832
                                                    100321583D
32   RT 15. 43
                              003226848
      30|c|72435| | | |l74490| |A|0.100J
                                              |003238870
      32RT17.85
                        00324988D
      30|c|877098||||373538||A|0.010J
                                                00325B9B1
      32|||RT|5.03|
                       00326C8D1
30c2051243    496702  A 0. 020
                                                       00327EAA7
      32   RT 22.52
                        00328FAA6
3.10  Record Type  20s  and 30s for  Pesticides  (Col 1) —  Initial Calibration
      Single Component — Individual Standard Mix B (QC Code 'CLM')
20 | ]INDBLIS|O|CLM| | I80oo|xi20i|92|o4|2o|i2|i3|
                                                            003292 199
21
         lNDBL15
                               2.0  0033034C3
      22||||||||i|||||l. 0)003314189


      30|c|319857| | | |301500| |A|o.010| | | || | I I I M I I I I I I |003326275
32   RT8.25
                             0033371A6
      30|c|319868| | | |946215| JA|0.010| || M M I M I I M M I |0033492C4
      32   RT 9.27
                       00335A1F8
30|c|7421934| | | |461669| |A|0.020| | |  | || M
                                                       00348EE52
      32   RT 15. 69
                       |00349FE6F

                               H-66
                                                                       OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
      30|c|5103719| | | |781319| |A|o.010| | | || I I I I I I I I I I I I (00350204E
32 | | |RT|l2.33J
                              003513041
30 |c| 5103742 | | | | 804407 | JA|0.010|
                                                       003525220
32 | | |RT|ll.87|
                             |00353622F
      30|c|877098| | | J373538| |AJ0.010| | | | | | | I I I I I I I I I I |003548353
32| | |RT|5.04|
                            |00355927E
30 G 2051243    496702 | |A| 0.020 |
                                                       00356B454
      32|||RT|22.53|
                       |00357C454
3.11  Record Type 20s  and  30s for Pesticides (Col 1) —  Instrument  Blank (QC
      Code 'LIB')

      20  PIBLKX1 1 LIB  18000 X1201 92 04 21 12 19  ML 1000 30 | | | 004663C44
      21
      22
         PIBLKXl
                               2.0  004674F64
                  10000 1.0    004686077
      3o|c|3i9846||UG/L|BDL|||  |u|o.025|

      30|c|319857||UG/L|BDL|||||u|0.025J

      30|c|319868||UG/L|BDL|||||u|0.025|
                                                |00469809D

                                                |00470AOC5


                                                I00471COEF
30 | CJ 11096825 | |UG/L|BDL| | | | | U | 0. 50 |


30|CJ877098| |UG/L| |0.040| | | | | | |  | | | |


32| | |RTJ5.06| | | | |  | | | | | |004981803


30 |c| 2051243 | j UG/L | NRP | 0. 042 | P |  | | | |
      32RT22.54
                        005004E58
                                              | | | | | | | | | [00496E7D1


                                               PO.O   0049708D6
                                                   | P | 281.8 | | |004993E4D
                                     H-67
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
3.12  Record Type 20s  and  30s  for Pesticides (Col 1)  — Method Blank (QC Code
      ' LRB ' )

      20| |PBLK58|l|LRB| | 18000 | X1201 | 92 | 04 | 21 | 14 | 31 | |ML|lOOOJ30| | |0052620A8.


      21 | | | |s| | |FPBLKOl|92|04|l7| | | | |  J2.0J |0052739F9
22
                  10000 1 . 0    005284BOC
      27) I I I  I I I |FJ9l|ll|l4|09|l9|P1248Al|N|


      30|c|319846||UG/L|BDL|||||u|0.050||||


      30|c|319857||UG/L|BDL|||||u|0.050J |||
                                            |005308E9E


                                            |||||00531AEC2


                                            I I I I I00532CEE8
30|c|ll096825| |UG/L|BDLJ | | | |u|l.OJ


30 | C| 877098 | |UG/L| | 0. 19 | | A | 200 | | |  |


32 | |  |RTJ5.06|  | | |   | !  | | |005603D46


30 | c| 2051243 |  |UG/L| |0.2l| |A|200| |  |
      32|||RT|22.53|
                        005627468
                                             I I I I I I I I I | 00558082B


                                             | S | 94 | | | P | 5. 6 | | |005592E19




                                               S  100   P 0.0    005616468
3.13  Record Type 20s and  30s  for Pesticides  (Col 1) — Matrix Spike Sample (QC
      Code 'LF1')
20 | |X1201MS|
                        | 18000|xi20l|92 | 04 | 21 | 16 | 08 | |MLJ300|30| | | 00563A08B
      21    S   488920 | 92 | 04 1 17 | | 92 | 04 | 15  2 . 0  00564BFE5
22
23
                  5000 1 . 0    00565D2D6
                         |MB|92|04|2l|l4|3l|FPBLKOl|p|o| | | S | 5 | R | 5 | 005660949
27 | | | | | | | |F|9l|ll|l4|09|l9|P1248Al|N|


30|c|319846| |UG/L|BDL| | | | |u|o.083| | |  |
                                                   00567285F


                                                       005684889
                                     H-68
                                                                 OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
30|c|58899| |UG/L| |0.82| |A|l.667J | | | |p|49|
32) | |RT|8.0l|
                             00572B839
30|c|76448| |UG/L| |o.69| |AJ1.667| | | | | P | 41 |
32| | |RT|8.90|
                             00574E6D6
                                                        |p|l2.3| | |00571A918
                                                       |p|6.2| | J00573D7A3
      30Jc|50293||UG/L||1.5||A|1.667|||||p|90||DJ6|||||p|0.0|||005875913
      32J j JRT|l5.43|
                        00588691E
      30|c|72435| |UG/L|BDL| | | | |u|0.83|


      30 | G| 53494705 | | UG/L | LLS | 0. 094 | J |
                                               00589877C


                                               p6.400590AD22
32   RT 18. 14
                              00591BD2E
30|c|7421934J |UG/L|LLS|0.091jJ| | | |

32J||RT|l5.68J |j|||j|j||00593F1D2

3o|c|5i037i9J|UG/L|BDL 11J|u|o.083
                                                    P 2.2   00592E1B6
                                                        005941209
30|CJ877098| |UG/L| |o.33J |A|lOO|
      32|||RT|5.03||||
                       0060568AC
30|c| 2051243 | | UG/L | |0.30J JA|lOO|
                                       |s|99|
                                                             00604598C
                                                     |p|3.3J | | 006068FOA
      32RT22.52
                        006079F09
3.14  Record Type 20s and 30s for Pesticides (Col 1) — Matrix Spike Duplicate
      Sample (QC Code 'LF2')

      20||xi201MSD|l|LF2||18000|X1201|92|04|21|16|41||MLJ300|30J|J00608CBB4
      21||||s|||488921|92|04|17||92|04|15||2.0||00609EBOF

                                     H-69
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
      22|||||||||| 500011.0| | | |00610FEOO
23
                         |MB|92|04|2l|l4|3l|FPBLKOl|p|o| | | S | 5 | R| 5 | 006113473
27| | | | | | | |F|9l|ll|l4|09|l9JP1248Al|N|

30|c|319846| JUG/L|BDL| |  | |  |u|0.083| | | |
30|c|76448| JUG/LJ |0.70| JA|l.667|
      32RT8.91
                       0061911FO
      30|c|50293| |UG/L| |1.6| |A|1.667|
      32RT15.45
                        00633944F
      30|c|72435| | UG/L | BDL | | | | |u|o.83|
      30|c|53494705| | UG/L | LLS | 0 . 098 | J |
32| | |RT|l8.16|
                              00636E928
      30|c|7421934| | UG/L | LLS | 0. 096 | J |
32 | | |RT|l5.70|
                              006381DD7
                                                   006125389
                                                      |0061373B3
                                                                0061802BC
                                                 P 0.0   006328438
                                               00634B2AD
                                              |p|l2.2|||00635D91A
                                              p4.3   006370DCC
      30JC|5103719||UG/L|BDL|||||u|0.083|||||  ||||||||||006393ECE
      30|c|877098| |UG/L| |0.32| |A|lOO| | | | | | | | | |s|97| | |p|0.0| | |00649858A


      32|||RT|5.04|||||||||||0065094B5
      30|c|2051243| |UG/L| |0.30| JA|lOO|
                                               |PJ3.3|||00651BB13
      32RT22.53
                        00652CB13
                                     H-70
                                                                 OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
3.15  Record Type 20s and 30s for Pesticides (Col 1) — Continuing Performance
      Check — Individual Standard Mix A (QC Code 'CLE')
20 | |INDAM72|OJCLE| | 18000 |X1201 | 92 | 04 | 21 | 22 | 12 |
                                                            006882FOF
21
         lNDAM72
                               2.o  00689423C
22|
                     |4.0|006904FOF
      30Jc|319846| |NGJ |o.039| |A|o.040| | | | | | | |DJ2.5| | | | | | j | J006917382
      32|||RT|6.94|||
                       0069282C1
30 |C | 58899 | |NG| | 0.035 | | A JO. 040 |
      32|| JRT|8.03|
                       00694B666
                                      D 12 . 5
                                                           00693A743
30c72435  NG  0.438
                         0.400
                                            D -9 . 5
00707132E
      32RT17.86
                        00708234C
      30JC|877098J |NG| J0.037|JA|0.040|
      32)||RT|5.05|
                       00710570A
      30|c|2051243||NG||o.06l||A|o.080|
      32)j|RT|22.52
                        007128D17
                                       D 7.5
                                                     0070947DE
                                                      J007117D18
3.16  Record Type 20s and 30s for Pesticides (Col 1)  — Continuing Performance
      Check — Individual Standard Mix B (QC Code 'CLE')

      20  INDBM78 0 CLE  18000 X1201 92 04 21 22 51    13   00713B40B
21
         INDBM78
                               2.o  00714C749
      22
                4.0 00715D41C
30 c 319857  NG  0.036  A0.040
                                       D 10. 0
                                                            00716F95A
32   RT 8. 26
                             00717088C
                                     H-71
                                                                 OLM03.0

-------
Exhibit  H  — Appendix A
Format of  Records  for Specific Uses
 30|c|319868| | NG | |o.038| |A|0.040| | | | | | | |o|5.o| j | | | | | | |
 32J | |RTJ9.28|
                             007193C35
30JC|309002| |NG| JO. 038) |A|0.040|
      32|||RT|9.83J
                        007216FAC
                                        D5.o
                                                            007182DOO
                                                           007206076
30|c|5103742| | NG | |o.038| |AJ0.040|
      32RT11.88
                              007371E78
      30 C 877098  NG  0.042  A 0.040
      32RT5.06
                             00739529B
30 | C| 205 1243 | | NG | |0.067| |AJ0.080|
      32RT22.54
                        0074188CO
                                        |o|5.o|
                                             D -5.0
                                                            007360E68
                                                            00738436E
                                                             0074078B5
3.17  Record Type 20s and 30s for Pesticides (Col 1) — Continuing Performance
      Check — Performance Evaluation Mixture (QC Code 'CLE')

      20 j | PEM90 0 | CLE | | 18000 | X1201 | 92 | 04 | 23 1 13 | 47 | |  | | 12 | | | 00777E850
21
         PEM90
                             2.0  00778FA7A
30|CJ319846| | NG | |0.025| |AJ0.020| | | | | | | |o|-25.0| | | | | | | | |007792031


32| | |RTJ6.93| | | | | | | | | |  |007802F65


30|c|319857| | NG | |0.025| |AJ0.020| | ]R|99.4| | | |D|-25.0| | | | | |
                                                                  O0781551E
      32RT8.25
                       00782644F
30Jc|58899| |NG| (0.022J JA|0.020| | |RJ98.7|
                                                                 00783891F
      32j ||RT|8.03
                       007849842
      30JC|72208||NG||0.108||A|O.IOO|||RJ98.9||||o|-8.o|
                                                          007880CB5
      32j ||RT|l4.27|
                   |PB|20.0|24.0|||0078924C6

                               H-72
                                                                       OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
      30|c|50293||NG||0.20l||A|0.200|||R|95.9||||D|-O.S||||||||(007904895
      32   RTl5.45
                         PB 4.0 24.0   007915FE1
      30|c|72435||NG||0.518||A|0.500|||R|97.9||||o|-3.6|||||||||0079283EC
32| | |RT|l7.86|
                              00793940A
30|CJ877098| | NG | |0.044| |AJ0.040| | |RJ98.7| | | |o|-10.0|
                                                                  007982735
32| | |RTJ5.05|
                            |007993661
30|c|2051243| | NG | |o.04l| JA|0.040| | | | | | | |o|-2.5|
                                                             008005COO
      32   RT 22.53
                              008016COO
3.18  Record Type 20s and 30s  for  Pesticides  (Col 1) — Field Sample chosen for
      MS/MSD (QC Code 'LSD')

      20||xi20l|l|LSDJ|l8000|xi20l|92|04|23|l5|42||ML|400J34|||0080296B4


      211 I I Isl I |FX120l|92|04|l7| I 92 I 04 I 15 I I 2.0 I I00803B55E
      22
      23
                  50001.0    00804C84F
                    MB 92 04 21 14 31 FPBLK01 P | 2 |
                                                              00805FBAE
      27||||||||F|9l|ll|l4|09|l9|P1248Al|N|||||||||008061AC4
      30|c|319846||UG/L|BDL|||||u|0.062J

      30|c|319857||UG/L|BDL|||||u|0.062|

      3o|c|3i9868||UG/L|BDL|||||u|o.062|
                                                 008073AEB
                                                 008085B14
                                                I008097B3F
      30|c|72559| | UG/L | LLS | 0 . 014 | JP |
32 | | |RT|l3.03|
                              008176DC4
      3o|c|72208| |UG/L|BDL| | | | |u|o.i2| | | |

      30|c|33213659| |UG/L|BDL| | | | |u|o,12|
                                     H-73
                                            |p|27.3| | |008165DD3
                                               008188C04
                                                  00819ACEF
                                                                       OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
      30|CJ877098||UG/L||0.26J|A|200||||||||||s|5l|||P|O.O|||00846E6CB


      32|||RT|5.04|||||||||||00847F5F6


      30|c|2051243||UG/L||0.26||AJ200||||||||||S|53|||p|7.7||J008481C7E


      32 I I |RT|22.531 I I I I I I I I I I008492C7E
                                     H-74                              OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
4.0   PESTICIDES (COL. 2)

4.1   Record Type 10 for Pesticides (Col 2)

      10 1 1 1 1 1 |GC|OLMOI.OP| |TESLAB| 1 1 | esoooooi | INSTP2 | (TEST LABS INC. | |ooooo2A3i

4.2   Record Type 11 for Pesticides (Col 2)

      ll|DB-608| |0.53| | | | | | | | | | J000024AD7

4.3   Record Type 20s  and 30s for Pesticides (Col 2)  — Initial Calibration Mean
      Values (QC Code 'MNC')
20| | |O|MNC| |18000|X1201|
30 c 319846   AVG 1137869
32   RT 4. 56 RTF 4. 51 RTT 4. 61
                                      62   000036465
                                           00005A047
                                                       0000485E4
30
32
30
32
30
32
30
32 |
30
32
30
32
|cj 12674112) | | | 3 5883) | | | | j | || | |
| | j RT j 5 . 69 | RTF | 5 . 62 | RTT ) 5 . 76 j | |
|c|l2674112| | | |l614l| | | | | | | | | | |
| | |RT|4.70|RTF|4.63 |RTTJ4.77| | |
|c|l2674112| | j | 18842 | | | | | | | | | | |
| )RTJ6.5l|RTF)6.44|RTT|6.58| | |
C|l2674112| | ) |l5946| | | | | | | | | | |
| |RT|7.30|RTF|7.23)RTT|7.37 | | |
C|l2674112| | | |l6309J | ( | | | | | | | |
| |RT|8.0l|RTF|7.94|RTT|8.08| | |
C|lll04282| | | |9686| | | | | | | | | | | j
|| | RT | 3 . 89 | RTF | 3 . 82 | RTT | 3 . 96 | | |
| | | | | | | | ) |00054566A
| ) |l|0005571C3
| | | | | | | | | J000568ECC
| | |2|00057AA13
||||||||||00058C730
| | |3|00059E285
| | | | | | | | | |00060FFAE
| | |4)000611AEO
||||||||||0006237EF
| [ |5|00063533C
| | | | | | | | |000646F7E
| | |l|000658AB9
                               H-75
                                                                       OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
      30 | C 1 11104282 | | |  | 3312 | 1 || | I I 1 I I I I I I I M I I I I 00066A6BF
32   RT 3 . 49 RTF 3 . 42 RTT 3 . 56
                                          2 00067C1EF
      30|CJ877098| | |AVG|559374J | | | | JR|2l| | || I I N I I I M |001242054
32   RT 2. 87 RTF 2. 82 RTT 2. 92
30|c|2051243| | JAVG|594715| | | | | |RJ16|
                                           001253A9F
                                                      J001265C02
      32 I | |RT|l6.34|RTF|l6.24|RTT|l6.44| | | | |  |0012778FF

4.4   Record Type  20s  and  30s  for  Pesticides  (Col  2)  —  Resolution  Check
      Standard (QC Code 'LPC')
20
21
30|
32
30
32
30 |
32 |
30 |
32 |
30 |
32 |
30
32 |
| IRESCOI |O|LPC| | is
| | | | | | IRESCOI | llj)
|c|877098| | | | | | | | |
| | |RT|2.89| | | | | | | |
|c|959988| | | | | | | | |
| | JRT|8.7l| | | | | | | |
|CJ5103742| | | | | | | |
|| | RT | 8 . 89 | | | | | | | |
!C|72559||||||||||
|||RT|9.35||||||||
[c|6057l| | | | | | | | | |
|||RT|9.61||||||||
C|l031078| | | | | | | |
|||RT|13.17|||||||
                                                          0014186BD
                    I I I |2.0| |001429965

                    |R|IOO.O| M M 11 M M

                    | | |00144C495


                    M^.OIIIMMHIII

                    I I I00146EF1C
                    | | |001481B13


                    R|99.0J | | | | | |


                    | I (001504499
                            (001526EC2


                             h6.3||||||||


                            |(001549AB1

                                     H-76
                                                00143B560
                                               00145DFEA
                                                 001470BCE
                                              001493566
                                              |001515F90
                                          001538AB1
                                                                 OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
30|CJ72435||||||


32|||RT|13.35|||

30|c|53494705| j |


32| | |RT114.06| | |


30|c|2051243| | | |


32 I I |RT|l6.35| I I
                                               00155B5AA
                             (00156C5AA
                                                 |00157E365
                             I00158F36E
                              001601B2A
                                           001590B1D
4.5   Record Type 20s and 30s for Pesticides (Col 2} — Performance Evaluation
      Mixture (QC Code 'CLE')

      20| IPEMOllolcLEl I 180001x1201 92|04J20l06 26||||l2|  |001614104
      21
         PEM01
                       2.0  00162531C
30 c 319846  NG  0.02l  A0.020   R 99 .
                                            D -5 . 0
      32   RT 4.59
                       001648725
      30|c|319857||NG||0.02l||AJ0.020|||R|94.9||  |o|-5.0
      32|||RT|7.3l|
                       00166BB15
      30|c|58899||NG||0.020||AJ0.020|||R|97.8|  ||D|O.O
      32j j |RTJ5.47
                       00168EDC9
30|c|50293| |NG| |0.182| JA|0.200| | |R|96.3| | | |o|9.o|
32   RT 11. 68
                         PB 1.0 12.0   00175B476
30|c|72435J JNG| | 0.501 1 |A|0.500| | JRJ94.8| | | |D | -0.2 |
      32RT13.34
                        00177E85D
30 C|877098| | NG | |0.037J |AJ0.040| | |R|95.8| | | |DJ7.s|
      32    RT2.89
                      |001838971

                               H-77
                                                                J0016377E7
                                                           00165ABF2
                                                         00167DE8D
      30|c|72208| |NG |0.103| |A|O.IOO| | |R|95.6| I I |o|-3.o| I I I I I I I |0017261EB


      32| | |RT|l0.09|  | | | |PB|11.0|12.0| | |0017379E1
                                                               001749D4D
                                                                00176D85E
                                                                001827A3C
                                                                       OLM03.0

-------
Exhibit H  — Appendix A
Format of  Records  for Specific Uses


      30|c|2051243J JNGJ |0.039J |A|0.040| | | | | | | \D\2.S\ | | | | | | | |00184AE9E

      32|||RT|16.34|||||||||||00185BEAA

4.6   Record  Type  20s and  30s  for  Pesticides (Col  2)  — Initial Calibration
      Multicomponent Standard  (QC Code  'CLS')
20
21
30
32
30
32
30
32
30
32
30
32
30
32
30
32
AR122102 C
1
) CLS
|AR122102
C 11104282
RT
3.89
C 11104282
RT
3.49
C 11104282
RT
5.67
C 11104282
RT
7.05
C 11104282
RT
8.02
C 877098]
RT
2.86
C 2051243
RT
16.33
18000 X1201 92 04 20|07 31 ||7| J002128054
2.0J 002139525
| 9686 A 0.400 | | | || 00214B62E

1 00215C641
3312 A 0.400 |J 00216E70E

308

220

213

55937

5947

2 00217F71E
7 A 0.400 ) 002181808
3 002192825
7 A 0.400 | 0022048FE
4 002215916
4 A 0.400 | 0022279EE
5 0022389FB
4 A 0.040 | 00224AB44
00225BA76
15 A 0.040 | J00226DC65
00227EC66
4.7   Record Type  20s  and 30s for  Pesticides  (Col 2)  —  Initial Calibration
      Single Component — Individual Standard A (QC Code 'CLM')

      20||lNDAM16|o|CLM||18000|X1201|92|04|20|12|46|||)ll||]00362366C


      21|||||||lNDAM16||||||||(2.0||003634997



                                     H-78                              OLM03.0

-------
                                                 Exhibit H — Appendix A
                                     Format of Records for Specific Uses
22
                4.000364566A
30|c|319846| | | | 1128992 | JA|0.040| | |R|94.7|
32| | JRT|4.56|
                       0036687A3
30|c|58899| j | |l040769| |A|o.040J | JR|96.3|
32 | | |RT| 5.44)
                       00368B815
32) | |RT|5.8l|
                       00370E873
30|c|959988| | | |834736J JA|o.040| | |R|95.
32   RT 8.68
                       003721905
30|c)6057l| | | |842038| |AJ0.080| j  |R|95.7|
32|)|RT|9.58|
                       003744874
30 | c| 72208 | | | | 630027 | |AJ0.080| | | R| 98 .
32|||RT|10.07|
                        003767882
30|c|72548| | | |599176| |A|0.080| ) |R|92.6|
32   RT 11 . 29
                        00378A8E1
30 | C| 50293 | | | | 572605 | JA|0.080| | |RJ99.2
32   RT 11. 66
                        00380D91A
32| | JRT|l3.33|
                        00382093E
30JC|877098| | | |558724J |A|o.040J | |R|96.3|
32| | |RT|2.87|
                       0038439B8
30|c|2051243| | | | 595863 | JA|0.080|
32||JRT|16.33|
                       |003866BB1

                               H-79
                                                      003657868
                                                     00367A8DC
30|c|76448| | | |l070468| |A|o.040| | |RJ96.8| | | | | J| I I I ! I I |00369D943
                                                    |0037109C3
                                                   |003733932
                                                    003756886
                                                    0037798EO
                                                    00379C90E
30|c|72435J | j J237099| |A|0.400J | |R|95.5| 1 | j | || I I I I I I I 00381F94A
                                                     003832A85
                                                 003855BBO
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
4.8   Record Type 20s  and  30s  for Pesticides (Col 2) —  Instrument Blank (QC
      Code 'LIB')

      20 | |PIBLKB2|1|LIB| 1 18000 | X1201 | 92 | 04 | 21 1 12 1 19 | JML|lOOO|30J | | 004708102
21
22
         PIBLKB2
                               2.0  004719422
                 | 10000 | 1.0 | | | J00472A535
30|c|319846| |UG/L|BDLJ | | | |u|0.025|


30|C|319857| |UG/L|BDL| | | j |u|0.025|


30|c|319868| |UG/L|BDL| ( j | |u|o.025|
                                                      |00473C55B


                                                      |00474E583


                                                       0047505AD
30 | C | 11097691 | | UG/L | BDL | | | | | U | 0 . 50 |


30 |c | 11096825 | | UG/L | BDL | | | | | U j 0 . 50 )


3o|c|877098| |UG/L| |0.040| | | | | | | | | | |
32   RT 2 . 89
                       005025CC9
30|c|2051243| | UG/L | NRP | 0. Oil | P |
                                             | | | | | | | | | j | 004990B8E


                                             | | | | | | | | | | (005002C8F


                                             | |P|O.O| | (005014D94




                                             | | ( | | | |p|281.8| | |005038305
32) | |RT|l6.35|
                              005049312
4.9   Record Type 20s and 30s for Pesticides  (Col  2)  — Method Blank (QC Code
      ' LRB ' )

      20 | |PBLK58|1JLRB| | 18000 | X1201 | 92 | 04 | 21 | 14 | 31 | |ML| 1000 | 30 | | | 005306672


      21 | | | |s| | |FPBLK01J92|04|l7| | | | | |2.o| |005317FC3
22
      27
               ) | | 10000 | 1 . 0 | | j | 0053290D6
         |F|9l|ll|l4|09|l9|pl248Al|N|
                                                   00534D468
      30|c|319846| JUG/L|BDLJ ( | | |u|0.050| | || I I I I I I M I I I |00535F48C
                               H-80
                                                                       OLM03.0

-------
                                                       Exhibit H — Appendix A
                                           Format of Records for Specific Uses
      30|c|319857||UG/L|BDL|||||u|0.050|)|||||||||||||(0053614B2


      30|c|319868| |UG/LJBDL| | | | |u|0.050| || I I! M I II I I I I |0053734DA

      30|c|58899| |UG/L|BDL| | | | |u|o.050| | | | | | | | | | | | I I I |005385434
      30|c|ll096825| |UG/L|BDL| | | | |u|l.0| ! | | | || I I I I I I I! I 005624DF5
30 |c | 877098 | |UG/LJ |0.18| |AJ200|
32RT2.89
                             00564831B
30 | C| 2051243 | JUG/LJ |0.2l| |A|200|
                                             | S | 89 | | | P | 5. 6 | | J0056373E6
                                               S 100   P 0 . 0   00565AA3D
32   RT 16. 35
                              00566BA4A
4.10  Record Type 20s and  30s  for Pesticides — Matrix  Spike  Sample (QC Code
      'LF1' )
20 | |X1201MS|1JLF1| | 18000 | X1201 | 92 | 04 | 21 | 16 | 08 |
      21    s   488920 92 04 17  92 | 04 15  2 .0 |  0056805C7
                                                                 00567E66D
22
23
                  5000l.o    0056918B8
                         |MB|92|04|2l|l4|3l|FPBLKOl|p|o| | | S | 1 | R | 1 | 005704FOF
2?| | | | | | | |F|9l|ll|l4|09|l9|pl248Al|NJ


30|c|319846| | UG/L | BDL | | | | |u|o.083| | |  |


30|c|319857| |UG/L|BDL| | | | |UJ0.083| | |  |


30|c|319868| | UG/L | BDL | | | | |u|0.083| | |  |


30|c| 58899) |UG/L| |0.73| |AJ0.833| |  | | |p
                                                   005716E25
                                                       005728E4F
                                                       00573AE7B
                                                       00574CEA9
                                               |D|O| | | | | P | 12 . 3 | | | 00575F7AO
32 | | |RT | 5. 43)
                             0057606CE
30 | C| 76448 | |UG/L| |0.65| |A| 0.833 |
                                                                005772EF6
      32) | |RTJ5.8l|
                      |005783E26

                               H-81
                                                                       OLM03.0

-------
Exhibit H  — Appendix A
Format of  Records  for Specific Uses
      30Jc|50293| |UG/L| |l.s| JA|l.667| | | | |p|90| \D\6\ \ \ \ |P|O.O| | |00591C120


      32|||RT|ll.66||j |||||||[00592D12C


      30|c|72435| |UG/L|BDL| | | | |u|0.83| i i i | i | i I I i i I i I |00593EF8A


      30|c|53494705| |UG/LJLLS|0.10|J| | | | ! | | I I I I I I II lpl6-4! I (005941434

      32|||RT|l4.03||||||||||(005952430


      30|c|7421934||UG/L|L


      LSJ0.089|j| | | I II I 1 I I I I I I I lpl2-2! I |0059648C9

      32|||RT|l2.34|||  {(I  |||0059758C7
      30|c|877098| JUG/L| |o.32| |AJ100| | | | | | | | I Is!97! I I?!3-1! I |00608C07E

      32|||RT|2.86||||||||j||00609CFBO

      30|CJ2051243||UG/L||0.31||A|100||||||||||S|94|||P|3.3||J00610F61C

      32) | |RT|l6.33| | | |  | || I I I  |00611061D

4.11  Record Type 20s and 30s for Pesticides — Matrix Spike  Duplicate Sample
      (QC Code  'LF2')

      20||xi201MSD|l|LF2l|18000|X1201|92|04|21|16|41|JML|300|30|||0061232C8

      21||||s|||488921|92|04|17||92|04|15||2.0||006135223

      22||||||||||5000|1.0||||006146514

      23|||||j||||||||||||MB|92|04|2l|l4|3l|FPBLKOl|p|o|)|S|1|R|1|006159B6B

      27||||||j  |F|9l|ll|l4|09|l9|pl248Al|NJ||||||||00616BA81


      30|c|319846| |UG/L|BDL| |  | | |u|o.083J | | | | I I I I M  I I I I |00617DAAB

                                     H-82                               OLM03.0

-------
32   RT 5 . 82
                       006238A90
30 | C| 50293 | |UG/L| | 1.6 | | A) 1.667 |
32   RT 11. 67
                        006371D9D
3o|c|72435J |UG/L|BDL| | | | |u|o.83 |

30|c|53494705| | UG/L | LLS | 0. 11 | J |  |
32   RT 14. 05
                        006407 171
30|c|7421934| | UG/L | LLS | 0. 092 | J |
32|j|RT|12.36|
                        00642A607
3o|c|5i037i9||UG/L|BDL|||||u|o.083|

30|CJ5103742| JUG/L|BDL|||||u|0.083|
30|c|877098| |UG/L| |0.32| |A|lOOJ
32 | | | RT | 2 . 87 | j | j j
                       006541CEC
30 | C| 2051243 | |UG/LJ | 0. 31 | | A| 100 |
32| | JRT|l6.35|
                        006565365
                                                 Exhibit H — Appendix A
                                     Format of Records for Specific Uses
30|c|76448| |UG/L| J0.66| |A|o.833J | | | |P| |79JD|l|
                                                          006227B5F
                                                  |0.0|| J006360D90
I|006383BFB


|p|l2.2||(006396169




|p|4.3|j|006419607




|||||00643C6FE


I I M I00644E7F1
                                      ||s|96|||P|0.0|||006530DB9
                                         S 94   P 3 . 3   006554358
                               H-83
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
4.12  Record Type 20s and 30s for Pesticides (Col 2)  — Calibration Verification
      Multicomponent Standard  (QC Code  'CLE')
20
21
30
32
30
32
30
32
30
32
30
32
30
32
30
32
AR122102JO
1
CLE 18000 (X1201
VAR1221V2 | (2.0
C 11104282
RT
3.89)
C 11104282
RT
3.49|
C 11104282
RT
5.67|
C 11104282
RT
7.05|
C 11104282
RT
8.02|
C 877098
RT
1
2.86|
C 2051243 |
RT|
16.33














| 9686 A 0.400
92|07 20 06 23 |7.| 1 002128054
002139525
| 00214B62E
| |l 00215C641
3312 A 0.400
00216E70E
2 00217F71E
3087 A 0.400
002181808
3 002192825
[2207 A 0.400
0022048FE
4 002215916
2134 A|0.400
0022279EE
S 0022389FB
559374 A 0.040
00224AB44
00225BA76
594715| A 0.040
00226DC65
00227EC66
4.13  Record Type 20s and 30s  for  Pesticides  (Col 2) — Field Sample chosen for
      MS/MSD (QC Code 'LSD')

      20|jX120l|l|LSD||18000 X1201|92|04j23|15|42j|MLJ400|34|||008061655


      211 | | |SJ | |FX120l|92J04J17| |92 | 04115| | 2 . 0 | |0080734FF


      22 I I I I I I I I I |5000|1.0|  | | |0080847FO


      23| | | | | | | | | | | |  | | | |  | |MB|92|04|2l|l4|3l|FPBLKOl|PJ2| | | | |  | J008097B4F



                                     H-84                               OLM03.0

-------
                                                 Exhibit H — Appendix A
                                     Format of Records  for Specific Uses
27
          F9llll409l9pl248AlN
                                             008109A65
30 | C | 319846 | | UG/L | BDL | | || I u I ° • °62 I I I I I I I I I I I I I I I I 00811BA8C

30|c|319857| |UG/L|BDL| | | | |u|o.062| | | | | | | | | | | | | |  |00812DAB5
3o|c|6057i||UG/L|BDL|||||u|o.i2|

30|c|72559||UG/L|LLS|0.Oil[JP|||
32|||RT|9.27|
                       00821EC9B
3o|c|72208||UG/L|BDL|||||u|o.i2
30|c|33213659||UG/L|BDL||||  UJ0.12|
30 C 72548  UG/L BDL   I  U 0.12
30 |C| 11096825 | |UG/L BDL | ||||u|


30|c|877098| |UG/L| |0.26J |A|200|
32| | |RT|2.86|
                      |00851720A
30|c|2051243| |UG/L| |0.28| |A|200|
32|||RT|l6.34|
                        00853A8AC
                                            I I |00819B90E


                                            |PJ27.3|||00820DD67




                                            I||008220ADB


                                                 |008232BC6


                                               008244A21
                                                 008493D1C
                                        S 51   po.0   0085062D8
                                         s55   p7.7   0085298AO
                               H-85
                                                                 OLM03.0

-------
Exhibit H — Appendix A
Format of Records for Specific Uses
5.0   TYPE 90 — COMMENTS RECORD

      This record could appear  anywhere  after the Record Type 10.  The comment
      on this record will usually  apply  to the most previous record type.   The
      example of Record Type 90 below would appear  after  the Record Type 10 -
      indicating the software and version number that was used to generate this
      data.
      90|CCS SYSTEMS SOFTWARE VERSION  3.5|000023A75
                                           U.S. Environmental Protection Agency
                                           Region 5, Library (PL-12J)
                                           77 West  Jackson Boulevard, 12th Floor
                                           Chicago,  IL  60604-3590
                                      H-86                               OLM03.0

-------