EPA 600/4-81-057
                 United States
                 Environmental Protection
                 Agency
&EPA         Research and
                 Development
                    The Analysis of Aromatic Chemicals ,
                   In Water by the Purge and Trap Method

                            Method 503.1
                 Prepared for
                     Joseph A. Cotruvo
                     Director
                     Criteria and Standards Division
                     Office of Drinking Water
                 Prepared by
                     Thomas A. Bellar
                     James J. Lichtenberg
                     Physical and Chemical Methods Branch
                     Environmental Monitoring and Support Laboratory
                     Cincinnati, Ohio  45268

-------
                      The Analysis of Aromatic Chemicals
                    in Water by the Purge and Trap Method
                                 Method 503.1

1.   Scope
    1.1   This method is applicable in the determination of those compounds,
         listed in Table I,  contained in finished drinking water, raw source
         water, or drinking  water in any stage of treatment.
    1.2   The actual detection limits are highly dependent upon the ability
         of the analyst to properly maintain the entire analytical system.
         Using carefully optimized equipment, the method has  been proven to
         be useful for the detection and measurement of multicomponent mix-
         tures spiked into finished water, carbon-filtered finished water,
         and raw source water at concentrations between 0.05  and 0.5 ug/1.
         The method as described is capable of accurately measuring those
         compounds listed in Table I over a concentration range of 0.05 to
         5.0 ug/1.  In addition, it is possible to measure individual com-
         pounds up to 1500 ug/1.  However the ability to measure complex
         mixtures containing co-eluting or partially resolved compounds with
         concentration differences larger than a factor of 10 is hampered.
  U.  S.  Environmental  Protection Agency,  Environmental  Monitoring and
  Support  Laboratory,  Physical  and Chemical  Methods Branch,  Organic Analyses
  Section,  Cincinnati,  Ohio 45268, May 1980.

-------
    1.3  This method is recommended for use only by analysts experienced in
         the measurement of purgeable organics at the low yg/1 level or by
         experienced technicians under the close supervision of a qualified
         analyst.
2.  Summary
    2.1  An extraction/concentration technique is incorporated within the
         method which enhances the quantities of certain compounds injected
         into the gas chromatograph by a factor of 1000 over direct injec-
         tion gas chromatography.
    2.2  Aromatics are extracted by an inert gas which is bubbled through
         the aqueous sample.   The  compounds, noted in Table I along with
         other organic constituents which exhibit low water solubility and
         boil less than 200°C, are efficiently transferred from the
         aqueous phase to  the gaseous phase.  These compounds are swept from
         the purging device and are trapped in a short column containing a
         carefully selected mixture of sorbent materials.  After a pre-
         determined period of time, the trapped components are dried then
         thermally desorbed and backflushed onto the head of a gas
         chromatographic column where they are separated under programmed
         temperature conditions.
    2.3  Measurement is accomplished with a photoionization detector which
         minimizes interference and baseline instability problems commonly
         encountered with  flame ionization detectors.
    2.4  If sufficient material is present, confirmatory analyses are or may
         be performed by gas  chromatography-mass spectrometry.

-------
2.5  Aqueous standards and samples are extracted and analyzed under
     identical conditions to compensate for extraction losses.
2.6  The total analysis time is approximately 1 hour per sample for all
     of the compounds listed in Table 1.
Interferences
3.1  Impurities contained in the purge gas and organic compounds out-
     gasing from the plumbing ahead of the trap usually account for the
     majority of contamination problems.  The presence of such inter-
     ferences is easily monitored using the quality control program
     described herein.  Sample blanks are normally run between each set
     of samples.  When a positive response is noted in the sample blank,
     the analyst should analyze a method blank in order to identify the
     source of contamination.  Method blanks are run by charging the
     purging device with reagent water and analyzing it in the normal
     manner.
     Whenever potential interfering peaks are noted in the method blank,
     the analyst should change the purge gas source and regenerate the
     molecular sieve purge gas filter.  Subtracting blank values from
     sample results is not recommended.  The use of non-TFE plastic
     tubing, non-TFE thread sealants, or flow controllers with rubber
     components in the purging device should be avoided since such
     materials out-gas organic compounds which will be concentrated in
     the trap during the purge operation.  Such out-gasing problems are
     common whenever new equipment is put into service.  With use, minor
     out-gasing problems generally cure themselves.

-------
3.2  Several instances of accidental sample contamination have been
     noted and attributed to diffusion of volatile organics through the
     septum seal and into the sample during shipment and storage.  The
     sample blank is used as a monitor for this problem.
3.3  For compounds that are not efficiently purged, such as naphthalene,
     small variations in sample volume, purge time, purge flow rate,
     purging device geometry, or purge temperature can affect the
     analytical result.  Therefore, samples and standards must be
     analyzed under identical conditions.
3.4  Cross-contamination can occur whenever high level and low level
     samples are sequentially analyzed.  To reduce the likelihood of
     this, the purging device and sample syringe should be rinsed out
     twice between samples with reagent water.  Whenever an unusually
     concentrated sample is encountered, it is necessary that it be
                                            *
     followed by a sample blank analysis to check for sample cross-
     contamination.  For samples containing large amounts of water
     soluble materials, suspended solids, high boiling compounds or high
     levels of compounds being determined,  it may be necessary to wash
     out the purging device with a  soap solution, rinse with distilled
     water, and then dry in an oven at 105°C between analyses.
3.5  Qualitative misidentifications are a potential problem in gas
     chromatographic analysis.  Whenever samples whose qualitative
     nature is unknown are analyzed, the following precautionary
     measures should be incorporated into the  analysis.
     3.5.1  Perform duplicate analyses using columns with dissimilar
            polarities.

-------
         3.5.2  Select a lamp for the photo ionization detector that provides
                the minimum ionization potential required to ionize the
                compounds of interest.  (See photoionization detector
                operators manual.)
         3.5.3  Whenever possible, use GC/MS techniques which provide
                unequivocal qualitative identifications.
4.  Apparatus
    4.1  The purge and trap equipment consists of three separate pieces of
         apparatus:  the purging device, trap, and desorber. Construction
         details for a purging device and an easily automated trap-desorber
         hybrid which has proven to be exceptionally efficient and repro-
         ducible are shown in Figures 1 through 6 and described in 4.1.1
         through 4.1.3.  All of the single lab data supplied in this method
         were obtained from an instrument of this design.
         4.1.1  Purging Device - Construction details are given in Figure 1
                for an all-glass 5 ml purging device.  The glass frit
                installed at the base of the sample chamber allows finely
                divided gas bubbles to pass through the sample while the
                sample is restrained above the frit.  Gaseous volumes above
                the sample are kept to a minimum to eliminate dead volume
                effects, yet allowing sufficient space for most foams to
                disperse.  The inlet and exit ports are constructed from
                heavy wall 1/4 inch glass tubing so that leak-free removable
                connections can be made using "finger-tight" compression
                fittings containing Teflon ferrules.  The removable foam
                trap is used to control samples that foam.

-------
4.1.2  Trapping Device - The trap (Figure 2) is a short gas chro-
       matographic column which at ambient temperature (22°C)
       retards the flow of the compounds of interest while venting
       the purge gas.  The trap is constructed with a low thermal
       mass so that it can be rapidly heated for efficient desorp-
       tion, then rapidly cooled to room temperature for recycling.
       Variations in the trap ID, wall thickness, sorbents, sorbent
       packing order, and sorbent mass can adversely affect the
       trapping and desorption efficiencies for certain compounds
       shown in Table I.  Pack the trap according to Figure 2.
4.1.3  Desorber assembly - Details for the desorber are shown in
       Figures 3 through 6.  With valve 1 in the Purge-Sorb
       position (See Figure 3), the effluent from the purging
       device passes through the trap where the movement of the
       organics is retarded.  The GC carrier gas also passes
       through valve 1 and is returned to the GC.  With valve 1  in
       the Purge-Sorb position, the operation of the GC is in no
       way impaired; therefore, routine liquid injection analyses
       can be performed using the gas chromatograph in this mode.
       After the sample is purged, residual water is removed from
       the trap by turning valve 2 to the "trap dry position" (See
       Figure 4).  The purge gas is rerouted so that it does not
       pass through the purging device.  At a flow rate of 40
       ml/minute the retention time for water and methanol on the
       23 cm Tenax trap is less than 3 minutes; therefore, with

-------
valve 2 in the "trap dry position" for four minutes most of
the residual water is removed from the trap by the anhydrous
purge gas.  (Both water and methanol cause a large negative
response early in the analysis.  This can cause measurement
errors unless this drying procedure is used.)  After the
trapped sample has been dried, valve 1 is turned to the
desorb position (See Figure 5).  In this configuration, the
trap is coupled in series with the gas chromatographic
column, allowing the carrier gas to backflush the trapped
materials onto the analytical column.  Just as valve 1 is
turned, the power is turned on to the resistance wire wrapped
around the trap.  The power is supplied by an electronic
temperature controller.  Using this device, the trap is
rapidly heated to 180°C with minimal temperature overshoot
and then maintained at 180°C.  The trapped compounds are
released as a "slug" to the gas chromatograph by this heat
and backflush step.  Normally, packed columns with theoreti-
cal efficiencies near 500 plates/foot under programmed
temperature conditions can accept such desorb injections
without altering peak geometry.
Substituting a non-controlled power supply, such as a
manually-operated variable transformer, will provide
non-reproducible retention times and poor quantitative data
unless the Injection Procedure in Section 8.4.2 is used.
After each sample has been injected into the gas chromato-
graph, the trap should be reconditioned.  Condition the

-------
            trap by turning valve 1 to the purge-sorb position and valve
            2 to the dry position.   Heat the trap to 180°C for 7
            minutes (See Figure 6).  Cool the trap to room temperature
            and turn valve 2 to the purge position before analyzing the
            next sample.
     4.1.4  Several Purge and Trap devices are now commercially avail-
            able.  It is recommended that the following be taken into
            consideration when purchasing a unit:
            a. Compatibility with the gas chromatograph to be used for
               the analysis.
            b. Includes a 5-ml purging device similar to that shown in
               Figure 1.
            c. The Tenax portion of the trap must meet or exceed the
               dimensions shown in Figure 2.
            d. Except for sample introduction, select a unit that has as
               many of the purge-trap functions automated as possible.
            e. The trap-dry mode (valve 2) will likely need to be added
               to commercial units.
4.2  Gas chromatograph - The gas chromatograph must be programmable and
     capable of operating at 40°C ± 1°C.  The gas chromatograph must
     be equipped with automatic flow controllers so that the column flow
     rate will remain constant throughout the program.  It may be neces-
     sary to cool the column oven to < 30°C.  (See Section 8.4.2.)
     Therefore, a subambient column controller may be required.
4.3  Gas chromatograph detector - A high temperature photoionization
     detector of new design equipped with a 10.2 eV lamp is used as a

-------
     semi-specific detector for those compounds described in Table 1.
     The unit is operated with an electrometer/lamp power supply.  The
     electrometer must be capable of stable, noise-free operation at
           -12
     1 x 10    amps with a full scale response time of < 1 second.
     The HNU Systems Inc., Model PI-52 photoionization and PI-52
     electrometer/Iamp power supply equipped with a 10.2 eV lamp has
     been found satisfactory for this purpose.
4.4  Gas Chromatographic Columns - Column 1 provides outstanding
     separations for a wide variety of aromatic hydrocarbons.  Column 1
     should be used as the primary analytical column because of its
     unique ability to resolve para, meta, and ortho aromatic isomers.
     Column 2, an extremely high polarity column, has been used a number
     of years for resolving aromatic hydrocarbons from alkanes in
     complex samples.  However, since the resolution between some of the
     aromatics is not as efficient as Column 1, it is recommended that
     Column 2 be used as a confirmatory column.
     4.4.1  Column 1 - Six feet long x 0.082 inch ID #304 stainless
            steel or glass tubing.  Packed with 5% SP-1200 + 1.75%
            Bentone 34 on 100/120 mesh Supelcoport.  The carrier gas is
            helium at a flow rate of 30 ml/minute.  The temperature
            program sequences are as follows:  For lower boiling
            compounds, operate at 50°C isothermal for 2 minutes then
            program at 6°/minute to 90°C and hold until all
            compounds have eluted.  For a higher boiling range of
            compounds, operate at 50°c isothermal for 2 minutes, then
            program at 3°/minute to 110°C and hold until all
            compounds have eluted.

-------
            NOTE:  In order to provide adequate resolution between the
            aromatic hydrocarbons at a flow rate compatible with the
            photoionization detector, it has been found that the 0.082
            ID column is necessary.  Variations in column ID and flow
            rate will sacrifice resolution or sensitivity.
            NOTE:  Whenever column 1 is not being used, maintain it at
            the upper temperature of the program, i.e. 90°C or
            110°C.  Condition new Bentone/SP-1200 columns at 120°C
            for several days with flow before connecting to the detec-
            tor.  See Figures 7 and 8 for sample chromatograms.
     4.4.2  Column 2 - Six feet long x 0.1 inch ID #304 stainless steel
            or glass tubing packed with 5% 1,2,3-tris (2-cyanoethoxy)
            propane on 60/80 mesh Chromosorb W-AW.  The carrier gas is
            helium at a flow rate of 30 ml/minute.  The temperature
            program sequence is as follows:  40°C isothermal for 2
            minutes then 2°/minute to 100°C and hold until all
            compounds have eluted.  See Figure 9 for an example
            chromatogram.
4.5  Sample containers - 40 ml screw-cap vials sealed with Teflon-faced
     silicone septa.
          Vials and caps - Pierce #13075 or equivalent
          Septa - Pierce #12722 or equivalent
4.6  Syringes - 5-ml  glass hypodermic with Luer-lok tip (2 each).
4.7  Micro syringes - 10, 100 ul.
4.8  Micro syringe - 25 yl with a 2" by 0.006 inch I.D. needle (Hamilton
     #702N or equivalent).

-------
    4.9  2-way syringe valve with Luer ends (3 each).
    4.10 Modified 500 and 1000 ml volumetric flasks.  See Figure 10.
5.  Reagents and Materials
    5.1  Trap Materials
         5.1.1  Porous polymer packing 60/80 mesh chromatographic grade
                Tenax GC (2,6-diphenylene oxide).
         5.1.2  OV-1 (3%) on Chromosorb-W 60/80 mesh.
    5.2  5% SP-1200/1.75% Bentone 34 on 100/120 mesh Supelcoport.
    5.3  5% 1,2,3-tris (2-cyanoethoxy) propane on 60/80 mesh Chromosorb W-AW.
    5.4  Dechlorinating compound-crystalline sodium thiosulfate, A.C.S.
         Reagent Grade.
    5.5  1:1 Hydrochloric Acid Solution in reagent water (must be shown to
         be interference-free).
    5.6  Activated carbon (for preparation of reagent water) - Filtrasorb-
         200, available from Calgon Corp., Pittsburgh, PA, or equivalent.
    5.7  Reagent water
         5.7.1  Reagent water is defined as water free of interference when
                employed in the purge and trap procedure described herein.
                It is generated by passing tap water through a carbon filter
                bed containing about 1 Ib. of activated carbon.
         5.7.2  A Millipore Super-Q Water System or its equivalent may be
                used to generate deionized reagent water.
         5.7.3  Reagent water may also be prepared by boiling water for 15
                minutes. Subsequently, while maintaining the temperture at
                90°C, bubble a contaminant-free inert gas through the water

-------
            for one hour.  While still hot, transfer the water to a
            narrow-mouth screw-cap bottle with a Teflon seal.
            NOTE:  Test reagent water daily by analyzing according to
            paragraph 8.
5.8  Standards
     5.8.1  Obtain 97% pure reagent grade reference standards.
5.9  Standard Stock Solutions.
     NOTE:  Because of the toxicity of some of the reference compounds,
     it is necessary to prepare primary dilutions in a hood.  It is
     further recommended that a NIOSH/MESA approved toxic gas respirator
     be used when the analyst handles high concentrations of such
     materials.
     5.9.1  Place about 9.8 ml of methyl alcohol into a 10 ml ground-
            glass stoppered volumetric flask.  Allow the flask to stand,
            unstoppered, for about 10 minutes or until all alcohol-
            wetted surfaces have dried and weigh to the nearest 0.1 mg.
     5.9.2  Using a 100 yl syringe, immediately add 2 drops of the
            reference standard to the flask, then reweigh.  Be sure that
            the 2 drops fall directly into the alcohol without contact-
            ing the neck of the flask.  Dilute to volume, stopper, then
            mix by inverting the flask several times.
     5.9.3  Calculate the concentration in micrograms per micro!iter
            from the net gain in weight.
     5.9.4  Transfer the standard solution to a 15 ml screw-cap bottle
            with a Teflon cap liner.

-------
     5.9.5  Store the solution at 4°C.
            NOTE:  Standard solutions prepared in methyl alcohol are
            stable up to 4 weeks when stored under these conditions.
            They should be discarded after that time has elapsed.
5.10 Calibration Standards
     5.10.1 In order to prepare accurate aqueous standard solutions, the
            following precautions must be observed.
            a. Do not inject more than 20 ul of alcoholic standards into
               100 ml of reagent water.
            b. Use a 25 ul Hamilton 702N microsyringe or equivalent.
               (Variations in needle geometry will adversely affect the
               ability to deliver reproducible volumes of methanolic
               standards into water.)
            c. Rapidly inject the alcoholic standard into the expanded
               area of the filled volumetric flask.  See Figure 10.
               Remove the needle as fast as possible after injection.
            d. Mix aqueous standards by inverting the flask three times
               only.
            e. Discard the contents contained in the neck of the flask.
               Fill the sample syringe from the standard solution con-
               tained in the expanded area of the flask as directed in
               Section 8.3.
            f. Never use pipets to dilute or transfer samples or aqueous
               standards.
            g. Aqueous standards are not stable and should be discarded
               after one hour unless preserved, stored, and sealed
               according to 6.3 and 6.4.

-------
         5.10.2 Prepare, from the standard stock solutions, secondary
                dilution mixtures in methyl alcohol so that a 20 ul injec-
                tion into 100, 500, or 1000 ml of reagent water will
                generate a calibration standard which produces a response
                close (±10%) to that of the unknowns.
         5.10.3 Purge and analyze the aqueous calibration standards in the
                same manner as the unknowns.
    5.11 Quality Control Check Standard (0.40 ug/1)
         5.11.1 From the standard stock solutions, prepare a secondary
                dilution in methyl alcohol containing 10 ng/ul of each
                compound normally monitored.  NOTE:  It may be necessary to
                prepare two or more quality control check standards so that
                all of the compounds in each mixture are adequately resolved
                for quantitative measurement.
         5.11.2 Daily, inject 20.0 yl of this mixture into 500 ml of reagent
                water and analyze according to the Procedure Section 8.
6.  Sample Collection and Handling
    6.1  The sample containers should have a total volume in excess of 40 ml.
         6.1.1  Narrow-mouth screw-cap bottles with the TFE fluorocarbon-
                faced silicone septa cap liners are strongly recommended.
                Crimp-seal serum vials with TFE fluorocarbon-faced septa are
                acceptable if the seal is properly made and maintained
                during shipment.
    6.2  Sample Bottle Preparation
         6.2.1  Wash all sample bottles and TFE seals in detergent.  Rinse
                with tap water and, finally, with distilled water.

-------
     6.2.2  Allow the bottles and seals to air dry at room temperature,
            place in a 105°C oven for one hour, then remove and allow
            to cool in an area known to be free of organics.
            NOTE:  Do not heat the TFE seals for extended periods of
            time (i.e., more than 1 hour) because the silicone layer
            slowly degrades at 105°C.
     6,2.3  When cool, seal the bottles with the TFE seals that will be
            used for sealing the samples.
6.3  Sample Preservation - It has been found that non-sterile samples
     containing aromatic hydrocarbons cannot be stored longer than 4
     hours because of biological degradation.  See Table 3.  Samples can
     be stabilized by adding free chlorine or by adjusting the pH to < 2
     with 1:1 hydrochloric acid.  See Table 4.  Free chlorine will react
     with styrene, and 2,3-benzofuran.  Therefore, if these compounds
     are to be determined in chlorinated water, it will be necessary to
     dechlorinate the sample with sodium thiosulfate at the rate of 1
     mg/ppm of free chlorine.  Once dechlorinated, the sample pH must be
     adjusted to < 2 with 1:1 hydrochloric acid.  If chemical preserva-
     tion is employed, the preservative is also added to the blanks.
     See Tables 2, 3, and 4 for recommended maximum holding times.
6.4  Sample Collection
     6.4.1  Collect a minimum of two replicates from each sample
            source.   Fill the sample bottles in such a manner that no
            air bubbles pass through the sample as the bottle is being
            filled.   Seal the bottles so that no air bubbles are

-------
       entrapped in  it.  Maintain the hermetic seal on the  sample
       bottle until time of analysis.
6.4.2  Sampling from a water tap.
       6.4.2.1 Turn on water and allow the system to flush.  When
               the temperature of the water has stabilized, adjust
               the flow to about 500 ml/minute and collect
               duplicate samples from the flowing stream.
6.4.3  Sampling from an open body of water.
       6.4.3.1 Fill a 1-quart wide-mouth bottle or 1-liter  beaker
               with sample from a representative area.  Carefully
               fill a minimum of two sample bottles from the
               sampling container as noted in 6.4.1.
6.4.4  If sodium thiosulfate preservative has been added to the
       sample bottles, then fill with sample just to overflowing,
       add two drops of 1:1 HC1 then seal the bottle, and shake
       vigorously for 1 minute.
6.4.5  Sealing practice for septum-seal screw-cap bottles.
       6.4.5.1 Open the bottle and fill to overflowing, place on a
               level surface, add 2 drops of 1:1 hydrochloric acid,
               position the TFE side of the septum seal upon the
               convex sample meniscus and seal the bottle by screw-
               ing the cap on tightly.
       6.4.5.2 Invert the sample and lightly tap the cap on a solid
               surface.  The absence of entrapped air indicates a
               successful seal.  If bubbles are present,  open the
               bottle, add a few additional drops of sample and
               reseal bottle as above.

-------
               NOTE:  If the septum seals are inverted (ie. the
               silicone side against the sample) then significant
               losses will occur.
               NOTE:  Be sure that the addition of 2 drops of 1:1
               HC1 will adjust the sample pH between 1 and 2.
               Check all new sample sources by adding 2 drops of
               1:1 HC1 to 40 ml of sample in a beaker.  Measure the
               pH with narrow range, (1.4 to 2.8), pH paper.
6.4.6  Blanks
       6.4.6.1 Sample blanks must be prepared and accompany the
               samples wherever the samples are shipped or stored.
               If the samples are immediately analyzed near the
               sampling site, blanks are not required.  Prepare
               blanks in replicate at the laboratory by filling and
               sealing a minimum of two sample bottles with pre-
               tested reagent water just prior to shipping the
               sample bottles to the sampling site.
       6.4.6.2 If the sample is to be preserved with sodium thio-
               sulfate, add an identical amount of preservative to
               the blanks.  Ship the blanks to and from the
               sampling site along with the sample bottles.  Open
               the blanks at the sample site, add 2 drops of 1:1
               HC1, then reseal.
       6.4.6.3 Store the blanks and the samples, collected from a
               given source (sample set), together.  A sample set
               is defined as all the samples collected from a given

-------
                        source (i.e., at a water treatment plant, the repli-
                        cate raw source waters, the replicate finished
                        waters, and the replicate blank samples comprise the
                        sample set).  Store the sample set in an area known
                        to be free of organic vapors.   See Table 2 for
                        maximum storage time.
7.  Conditioning Traps
    7.1  Condition newly packed traps overnight at 200°C by backflushing
         with an inert gas flow of at least 20 ml/min.  Vent the trap
         effluent to the room, not to the analytical column.
    7.2  Prior to daily use, condition traps 10 minutes while backflushing
         at 180°C.  The trap may be vented to the analytical column;
         however, after conditioning, the column must be programmed prior to
         use.
8.  Extraction and Analysis
    8.1  Adjust the purge gas (nitrogen or helium) flow rate to 40 ml/min.
         Attach the trap inlet to the purging device by turning valve 1 to
         the purge-sorb position (Figure 3).
    8.2  Open the syringe valve located on the purging device sample intro-
         duction needle (Figure 1).  Remove the plungers from two 5-ml
         syringes and attach a closed syringe valve to each.
    8.3  Open the sample bottle (or standard) and carefully pour the sample
         into one of the syringe barrels until it overflows.  Replace the
         syringe plunger and compress the sample.  Open the syringe valve
         and vent any residual air while adjusting the sample volume to 5.0
         ml.  Close the valve.  Fill the second syringe in an identical

-------
     manner from the same sample bottle.  This second syringe is
     reserved for a replicate analysis, if necessary.
8.4  Attach the syringe-valve assembly to the syringe valve on the
     purging device.  Open the syringe valve and inject the sample into
     the purging chamber.  Close both valves.  Purge the sample for 12.0
     ± .05 minutes.  After the 12-minute purge time, turn valve 2
     (Figure 4) to the dry position for 4 minutes.  Empty the purging
     device using the sample introduction syringe, follow with two 5-ml
     flushes of reagent water.  Leave the syringe valve open.  After the
     4-minute dry purge time, attach the trap to the chromatograph (turn
     valve 1 to the desorb position, Figure 5) and introduce the trapped
     materials to the GC column by rapidly heating the trap to 180°C
     while backflushing the trap with an inert gas between 20 and 60
     ml/min for 4 minutes.
     8.4.1  If the trap can be rapidly heated to 180°C and maintained
            at this temperature, the GC analysis can begin as the sample
            is desorbed, i.e., the column is at the initial 50°C
            operating temperature.  Start the temperature program and
            collect retention data just after valve 1 is turned.  The
            equipment described in Figures 3 through 6 will perform
            accordingly.
     8.4.2  With other types of equipment (see 4.1.4) where the trap is
            not rapidly heated or is not heated in a reproducible
            manner,  it is necessary to transfer the contents of the trap
            into the analytical  column at 30°C where it is once again
            trapped.   Once the transfer is complete (4 minutes), the

-------
            column is rapidly heated to the initial operating tempera-
            ture for analysis.  Start to collect retention data just as
            the column is heated.
            NOTE:  In severe cases, it may be necessary to cool the
            column down to 0°C.
     8.4.3  If injection procedure 8.4.1 is used and the early eluting
            peaks in the resulting chromatogram have poor geometry or
            variable retention times, then method 8.4.2 should be used.
8.5  After desorbing the sample for four minutes, recondition the trap
     by returning valve 1 to the sorb position.  See Figure 6.  Maintain
     the trap temperature at 180°C.  After approximately seven
     minutes, return valve 2 to the purge position and turn off the trap
     power.  Leave the syringe valve open.
8.6  Treat and analyze each sample and sample blank from the sample set
     in an identical manner (see 6.4.6.2) on the same day.
8.7  Prepare single point standards from the standard stock solutions
     (5.11) in reagent water that are close to the unknown in composi-
     tion and concentration (9.1).  The concentrations should be such
     that no more than 20 pi of the secondary dilution need be added to
     100 to 1000 ml of reagent water to produce a standard at the same
     level as the unknown.
8.8  As an alternative to single point calibration, Section 8.7, once
     the stability of the entire system is established, construct a
     calibration curve for each compound normally monitored over a
     concentration range that will bracket each sample.  Daily check the
     validity of this calibration curve using the 0.4 ug/1 quality
     control check sample.  The value obtained must be within ± 0.08
     ug/1.  If it is not, generate a new calibration curve or use 8.7.

-------
9.  Analytical Quality Control
    9.1  Analyze the 0.40 ug/1 quality control check sample daily before any
         samples are analyzed.  Instrument status checks and lower limit of
         detection estimations, based upon response factor calculations at
         five times the noise level, are obtained from these data.  In
         addition, response factor data obtained from the 0.40 ug/1  quality
         control check standard can be used to estimate the concentration of
         the unknowns.  From this information, the appropriate standard
         dilutions can be determined.
    9.2  Analyze the EMSL-Cincinnati volatile organics quality control
         samples or their equivalent on a quarterly basis.  The values must
         be within 20% of the true value.
    9.3  Analyze the sample blank or a method blank to monitor for potential
         interferences as described in Sections 3.1, 3.2 and 3.4.
    9.4  Daily perform the following instrument status checks, using the
         data gathered from blanks, duplicate analyses, and the quality
         check sample.
         9.4.1  Peak Geometry Check
                9.4.1.1 All of the peaks contained in the quality control
                        sample chromatogram must be sharp and symmetrical.
                        Peak tailing in excess of that shown in the  method
                        chromatograms must be corrected.  If only the com-
                        pounds eluting before ethylbenzene give random
                        responses, unusually wide peak widths, unstable
                        retention times, poor resolution, or are missing,
                        the problem is usually traceable to the trap/

-------
                        desorber.  See Sections 4.1.4 and 8.4.2.   If
                        negative peaks appear early in the chromatogram,
                        increase the dry purge time to 5 minutes.   Retention
                        times for the compounds should remain constant (less
                        than 10% variance)  throughout the day.
         9.4.2  Check the precision between replicate analyses.   A properly
                operating instrument should perform with an average relative
                standard deviation of less  than 6% over a concentration
                range of 0.1 to 100 ug/1.   Poor precision is generally
                traceable to:
                a.  Pneumatic leaks especially around the purging  device,
                   trap, and column.
                b.  Too high lamp intensity  power.   Should adjust  to about 50
                   to 60% for the 10.2 eV  lamp.
         9.4.3  The method blank analysis should represent less than a 0.1
                ug/1 response or less than  a 10% interference for  those
                compounds that occur routinely.
    9.5  Any instrument not performing according to 9.4 specifications
         should be  considered "out of control."  The system must  be "in
         control" before acceptable data can be generated.
10.  Calculations
    10.1  Determine  the concentration of the unknowns by comparing  the peak
         height (area) of the unknowns to that of the standard peak height
         (area) (8.7).  Round off the result to the nearest .01 yg/1 or two
         significant figures.

-------
                peak height sample
              = peak height standard x (concn'  std>
    10.2 Report the results obtained from the EMSL Quality Control
         Sample and the lower limit of detection estimates along with
         the data for the unknown samples.
    10.3 Calculate the limit of detection (LOO) for  each compound not
         detected using the following criteria:
                          (A x ATT)
         LOD(ug/D=0.4 (F77fF)
         where:    A = 5 times the noise level  in (mm)  at the exact
                   retention time of the compound or the baseline
                   displacement in (mm) from the theoretical  zero at
                   the exact retention time of  the compound.
11.  Accuracy and Precision
    11.1 Precision and Accuracy for Purge and Trap Method Using the Photo-
         ionization Detector under the conditions described in Section 4.3.
         11.1.1  Both Ohio River water and carbon-filtered tap water were
                 spiked with known amounts of selected  compounds.  The
                 spiked solutions were then sealed in septum-seal vials then
                 stored on the bench top for up to four weeks.  Samples were
                 randomly analyzed on several occasions.  Tables 2, 3, and 4
                 show the accuracy, precision,  and maximum holding  time data
                 obtained from this study.

-------
                          TABLE 1.  RETENTION DATA
                                  Column 1
Column 2   Lower Limit

Compound
Benzene
1,1,2-Trichloroethylene
a,-Trifluorotoluene
Toluene
1,1,2,2-Tetrachloroethylene
Ethyl benzene
1-Chlorocyclohexene-l
p-Xylene
Chlorobenzene
m-Xylene
o-Xylene
Iso-propylbenzene
Styrene
p-Bromofluorobenzene
n-Propyl benzene
tert-Buty 1 benzene
o-Chlorotoluene
p-Chlorotoluene
Bromobenzene
sec-Butyl benzene
1, 3, 5-Trimethyl benzene
p-Cymene
1,2,4-Trimethylbenzene
p-Di ch 1 orobenzene
m-Di Chlorobenzene
Cyclopropyl benzene
n-Butyl benzene
2,3-Benzofuran
o-D ichl orobenzene
Hexachlorobutadiene
1, 2, 4-Trichl orobenzene
Naphthalene
1,2,3-Trichlorobenzene
Program
A
199
223
275
340
360
491
518
518
542
542
574
595
544
664
681
786
804
804
804
829
851
909
909
999
1082
1082
1082
1283
1528
2035
2690
4280
4526
Program
B
199
231
296
384
406
606
637
653
689
689
738
768
834
852
879
975
985
990
999
1027
1043
1090
1090
1152
1211
1211
1211
1320
1425
1650
1928
2545
2631


165
142
168
255
168
375
345
403
481
403
518
455
690
740
518
595
681
—
807
595
612
681
750
975
901
...
765
1460
1161
1011
1535
2298
1820
of Detection
ug/l*
0.02
0.01
0.02
0.02
0.01
0.002
0.008
0.002
0.004
0.004
0.004
0.005
0.008

0.009
0.006
0.008
_ —
0.002
0.02
0.003
0.009
0.006
0.006
0.006

0.02
0.03
0.02
0.02
0.03
0.04
0.03
Column 1-6 Teet long x 3.082 ID stainless steel packed with 5%
SP-1200/1.75% Bentone 34 on 100/120 mesh Supelcoport.
   Program A:  50° C hold 2 minutes, 6°/min to 90° C
   Program B:  50° C hold 2 minutes, 3°/min to 110° C

Column 2-8 feet long x 0.10 inch ID stainless steel packed with 5%
1,2,3-Tris (2-Cyanoethoxy) propane on Chromosorb-W AW.
   Program:  40° C hold 2 minutes, 2°/minute to 100° C.
*Lower limit of detection -   99% confidence that the value is not zero
                              calculated from 7 runs at 0.04 ug/l.

-------
        Single Laboratory Accuracy and Precision for Aged Samples Containing Aromatic Compounds
                   and Selected Organohalides Spiked Into Chlorinated Drinking Water


Compound

Benzene
Trichloroethylene
a-Trichlorotoluene
Toluene
Tetrachloroethylene
Ethylbenzene
1-Chlorocyclohexene-l
p-Xylene
Chlorobenzene
m-Xylene
o-Xylene
iso-Propylbenzene
Styrene
p-Bromof luorobenzene
m-Propyl benzene
t-Butylbenzene
o-Chlorotoluene
p-Chlorotoluene
Bromobenzene
sec-Butylbenzene
1,3,5-Trimethylbenzene
p-Cymene
1,2,4-Trimethylbenzene
p-Dichlorobenzene
m-Di chlorobenzene
Cycloprepyl benzene
m-Butylbenzene
2,3-Dibenzofuran
o-Di chlorobenzene
Hexachlorobutadiene-1 ,
1 ,2,4-Trichlorobenzene
Naphthalene
1 ,2,3-Trichlorobenzene

Spike
Level
(yg/D
0.40
0.50
0.50
0.40
0.50
0.40
0.50
0.40
0.50
0.40
0.40
0.40
0.40
No data
0.40
0.40
No data
0.50
0.50
0.40
0.50
0.80
0.40
0.50
0.50
No data
0.40
0.40
0.50
3 0.50
0.50
0.50
0.50

Day 1
Recovery
~T£)
100
103
100
100
108
103
96
95
96
95
93
93
0.00

90
95

94
96
85
96
92
83
106
96

90
14
102
88
94
108
100

Samples
Analyzed3

7
10
9
7
10
7
10
7
10
7
7
7
7

7
7

8
10
7
10
5
7
10
10

7
7
9
10
10
8
10
-- ,- . .. 	 	 y ,,.„-,— ..

Mean
Recovery
r£7~
100
104
89
93
104
93
91
85
96
90
90
88
0.00

83
88

93
93
80
92
88
75
100
92

78
0.0
92
74
88
96
85

Standard
Deviation

0.082
0.037
0.048
0.022
0.040
0.032
0.029
0.029
0.029
0.028
0.026
0.030


0.030
0.030

.022
.030
0.034
0.040
0.012
0.029
0.029
0.040

0.049
-
0.033
0.062
0.047
0.062
0.046
Relative
Standard
Deviation
(%)
2.1
7.0
11.0
5.7
7.7
8.5
6.4
8.7
6.1
7.7
7.2
8.7


9.3
8.7

4.9
6.4
11. 0
8.7
2.8
8.7
8.7
8.7

16
-
7.1
17
11
13
11
Length
of
Study
T3ays)
15
28
28
15
28
15
28
15
28
15
15
15
0*

15
15

13*
28
15
28
6*
15
28
28

15
0*
28
28
28
20*
28
 ^f UMII1-' | *•> .J I MlfVJVSIIICr V*liV»l^fc*-»«  w » « • -w M j . • •

*Maximum recommended holding time.

-------
                                        TABLE 3

Single Laboratory Accuracy and Precision for Aged Samples Containing Aromatic Compounds
                and Selected Organohalides Spiked Into Ohio River Water


Compound

Benzene
Trichloroethylene
a-Trichlorotoluene
Toluene
Tetrachloroethylene
Ethylbenzene
1-Chlorocyclohexene-l
p-Xylene
Chlorobenzene
m-Xylene
o-Xylene
iso-Propylbenzene
Styrene
p-Bromof luorobenzene No
m-Propylbenzene
t-Butylbenzene
o-Chlorotoluene No
p-Chlorotoluene
Bromobenzene
sec-Butylbenzene
1 ,3,5-Trimethylbenzene
p-Cyrnene
1,2,4-Trimethylbenzene
p-Di chlorobenzene
m-Dichlorobenzene
Cyclopropy Ibenzene No
n-Butylbenzene
2,3-Dibenzofuran
o-Dichlorobenzene
Hexachlorobutadiene-1,3
1,2,4-Trichlorobenzene
Naphthalene
1 ,2,3-Trichlorobenzene

Spike
Level
(ygTTT
0.40
0.50
0.50
0.40
0.50
0.40
0.50
0.40
0.50
0.40
0.40
0.40
0.40
data
0.40
0.40
data
0.50
0.50
0.40
0.50
0.50
0.40
0.50
0.50
data
0.40
0.40
0.50
0.50
0.50
0.50
0.50

Day 1
Recovery
~T*1
12
86
74
8
85
10
86
12
23
17
12
0.0
0.0

15
43

45
17
10
16
34
5
90
96

3
0.0
.85
84
88
38
97

Samples
Analyzed3


6
6

6

6
















6
6



5
6
6
6
5

Mean
Recovery
(%)

87
87

83

86
















87
88



88
77
88
42
91

Standard
Deviation


0.015
0.081

0.014

0.034
















0.048
0.036



0.53
0.099
0.072
.20
.070
Relative
Standard
Deviation
(%)

3.5
23.

3.3

7.8
















10.9
8.3



12
26
16
98
15
Length
of
Study
(days)

18
18

18

18
















18
18



18
18
18
18
4
 randomTv analyzed tnrouonout tne studv oeriod.

-------
             Single Laboratory Accuracy and Precision for Aged Samples Containing Aromatic Compounds
                             and Selected Organohalides Spiked Into Ohio River Water


Compound

Benzene
Trichloroethylene
a-Trichlorotoluene
Toluene
Tetrachloroethylene
Ethylbenzene
1-Chlorocyclohexene-l
p-Xylene
Chlorobenzene
m-Xylene
o-Xylene
iso-Propyl benzene
Styrene
p-Bromof luorobenzene
m-Propylbenzene
t-Butylbenzene
o-Chlorotoluene
p-Chlorotoluene
Bromobenzene
sec-Butylbenzene
1,3,5-Trimethylbenzene
p-Cymene
1,2,4-Trimethylbenzene
p-Di chlorobenzene
m-Di chlorobenzene
Cycloprepyl benzene
n-Butylbenzene
2,3-Dibenzofuran
o-Dichlorobenzene
Hexachlorobutadiene-1,3
1,2,4-Trichlorobenzene
Naphthalene
1,2,3-Trichlorobenzene

Spike
Level
WT)
0.40
0.50
0.50
0.40
0.50
0.40
0.50
0.40
0.50
0.40
0.40
0.40
0.40
No data
0.40
0.40
No data
0.50
0.50
0.40
0.50
0.50
0.40
0.50
0.50
No data
0.40
0.40
0.50
0.50
0.50
0.50
0.50

Day 1
Recovery
"OH
100
104
84
98
100
95a
82
88 a
90
93a
95a
93a
83a

88a
90a

90
90
88 a
88
88
85a
94
96

85a
95a
96
82
88
82
86

Samples
Analyzed**

6
9
9
6
9

9

9








9
9




9
9



9

8
8
8

Mean
Recovery
~W
100
89
86
95
90

86

95








90
92




90
89



92

83
87
85

Standard
Deviation

0.014
0.028
0.037
0.028
0.035

0.034

0.027








0.023
0.027




0.032
0.037



0.033

0.038
0.075
0.042
Relative
Standard
Deviation
(%)
3.5
5.8
8.2
7.6
7.6

7.9

5.8








5.1
5.8




7.2
8.4



7.1

9.1
17.
10.
Length
of
Study
(days)
14
26
26
14
26
14
26

26
14
14
14
14

14
14

26
26
14
14
14
14
26
26

14
14
26
26
26
26
26

Average
Decay Rate
(%/day)





1.4

2.3

1.6
1.6
1.5
4.2

2.2
10



2.2
2.1
2.3
3.0



3.2
2.4

3.2



aSample 2 days old.
bSamples randomly analyzed throughout the study period.

-------
              OPTIONAL
              FOAM TRAP
1 4 IN. O.O. EXIT
          10MM. GLASS FRIT
          MEDIUM POROSITY
- EXIT 1/4
  IN. O.D.

— I4MM. O.D.

 INLET 1/4
 IN. O.D.
                                     SAMPLE INLET

                                       2-WAY SYRINGE VALVE
                                     •  17CM. 20  GAUGE SYRINGE NEEDLE
                                     6MM.O.O. RUBBER SEPTUM
                                         10MM. O.O.
                                           INLET
                                           1/4 IN. O.D.
                         1/16 IN. O.D.
                         STAINLESS STEEL
                                                    Tf
                                                        13X MOLECULAR
                                                        SIEVE PURGE
                                                        GAS RLTER
                                                           PURGE GAS
                                                           FLOW CONTROL
                   FIGURE 1.  PURGING DEVICE

-------
          PACKING  PROCEDURE
CONSTRUCTION
GLASS WOOL  5MM
  TENAX
           23CM
  3% OV-1     KM
  GLASS WOOL
             5MM
                                                           COMPRESSION FITTING NUT
                                                           AND FERRULES
                                                                     MFT. 7-VFOOT RESISTANCE
                                                                     WIRE WRAPPED SOLID
                                                                 THERMOCOUPLE/CONTROLLER
                                                                 SENSOR
                        ELECTRONIC

                        TEMPERATURE

                        CONTROL

                        AND

                        PYROMETER
                                                             TUBING  25CM 0.105 IN. I.D.
                                                             0.175 IK O.D. STAINLESS STEEL
                  TRAP INLET
           FIGURE   2   TRAP

-------
       CARRIER GAS FLOW CONTROL

    PRESSURE REGULATOR
LIQUID INJECTION PORTS
  PURGE GAS
  FLOW CONTROL \
13X MOLECULAR
SIEVE FILTER
           VALVE-3
           OPTIONAL 4-PORT COLUMN
           SELECTION VALVE
              TRAP INLET(TENAX END)
   VALVE-1   /   RESISTANCE WIRE
                                                         TRAP
                                                         22°C
                            COLUMN OVEN

                              CONFIRMATORY COLUMN
                            TO DETECTOR

                                ANALYTICAL COLUMN
                                                                           HEATER CONTROL
                                VALVE-2
                                                                NOTE:  ALL LINES BETWEEN
                                                                      TRAP AND GC
                                                                      SHOULD BE HEATED
                                                                      TO 80°C
             FIGURE  3   PURGE-TRAP SYSTEM  (PURGE-SORB MODE)

-------
        CARRIER GAS FLOW CONTROL

     PRESSURE REGULATOR
LIQUID INJECTION PORTS
  PURGE GAS
  FLOW CONTROL \
13X MOLECULAR
SIEVE FILTER
                            COLUMN OVEN


                              CONFIRMATORY COLUMN
                            TO DETECTOR

                                ANALYTICAL COLUMN
           VALVE-3
           OPTIONAL 4-PORT COLUMN
           SELECTION VALVE
              TRAP INLET (TENAX END)
  VALVE-1   /  x RESISTANCE WIRE
                                                                            HEATER CONTROL
                                                                NOTE: ALL LINES BETWEEN
                                                                      TRAP AND  GC
                                                                      SHOULD BE HEATED
                                                                      TO 80°C
                                                                                                       I
                                                                                                       •i
                                VALVE-2
              FIGURE 4     PURGE-TRAP SYSTEM    (TRAP-DRY MODE)

-------
       CARRIER GAS FLOW CONTROL

     PRESSURE REGULATOR
LIQUID INJECTION PORTS
  PURGE GAS
  FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
                            COLUMN OVEN


                              CONFIRMATORY COLUMN
                            TO DETECTOR

                                ANALYTICAL COLUMN
            VALVE-3
            OPTIONAL 4-PORT COLUMN
            SELECTION VALVE
              TRAP INLET (TENAX  END)
                  RESISTANCE WIRE
                                                                            HEATER CONTROL
                                                                NOTE: ALL LINES BETWEEN
                                                                      TRAP AND  GC
                                                                      SHOULD BE HEATED
                                                                      TO 80° C
                                VALVE-2
             FIGURE  5    PURGE-TRAP SYSTEM  (DESORB MODE)

-------
        CARRIER GAS FLOW CONTROL

     PRESSURE REGULATOR
LIQUID INJECTION PORTS
  PURGE GAS
  FLOW CONTROL \
13X MOLECULAR
SIEVE FILTER
            VALVE-3
            OPTIONAL 4-PORT COLUMN
            SELECTION VALVE
              TRAP INLET (TENAX END)
    VALVE-1  /   RESISTANCE WIRE
                         PURGING
                         DEVICE
                                          TRAP
                                          FLOW
                     TRAP
                     180°C
                            COLUMN OVEN


                               CONFIRMATORY COLUMN

                            TO DETECTOR

                                ANALYTICAL COLUMN
                                                                             HEATER CONTROL
                                 VALVE-2
                                                                 NOTE:  ALL LINES BETWEEN
                                                                       TRAP AND  GC
                                                                       SHOULD BE HEATED
                                                                       TO 80°C
              FIGURE  6    PURGE-TRAP SYSTEM   (TRAP-CONDITION MODE)

-------
               _- fc -.tfc—,
                            FULL SCALE RESPONSE 16*10^ AMPS
 C
 3J
 m
 O
 z
 a
 o
 2
 >

 o
 o
O
•n

m
tn
X
H

as
m
                                                1.1,2-TRICHLOROETHYLENE

                                 o-TRIFLUOROTOLUENE
                                    1,1,2,2-TETR ACHLOROETHYLENE
                                                  CHLOROBEN2ENE
o^CHLORQTOLUENE

             BROMOBENZENE

 1,3,5-TRIMETHYLBENZENE



    P-DICHLOROBEN2ENE

      m-DICHLOROBENZENE
                                             o-DICHLOROBENZENE
                          1 -CHLOROCYCLOHEXENE
                               HEXACHLOROBUTADIENE-i ,3
1,2.4-TRICHLOROBENZENE
                                  NAPHTHALENE


                               1.2.3-TRICHLOROBEN2ENE
                                ISIS
                               3)  S
                               m  e
s*
 0
n

-------
                                                             COLUMN: 5% SP-1200/1.76% BENTONE 34
                                                             PROGRAM: 60°C - 2 MINUTES 6°C/min. to 90°C
                                                             DETECTOR: PHOTOIONIZATION
                                                             SAMPLE: 0.40/ig/l STANDARD MIXTURE
(A
Z
8i|
Si
  «0
D
u.
                                                                                          20
22
                                        RETENTION TIME, minutes
                            FIGURE 8. CHROMATOGRAM OF TEST MIXTURE
                             •i

-------
               COLUMN: 5% 1.2.3-TRIS (2-CYANOETHOXY)
               PROPANE ON CHROMOSORB—W
               PROGRAM: 40°C-2minutes 2°C/min. to 100°C
               DETECTOR: PHOTOIONIZATION
               SAMPLE: 2.0 Mg/l STANDARD MIXTURE
             8        12
            RENTENTION TIME,minutes
FIGURE 9. CHROMATOGRAM OF TEST MIXTURE

-------
                                        6MM O.D HALF-HOLE
                                        CYLINDRICAL SEPTUM
                                       8MM O.D TUBING
                                       9MM. LONG
FIGURE 10.   MODIFIED VOLUMETRIC FLASK
                                           «-. —it,-?! Protection Agency

-------