DETERMINATION OF  ROTENONE  IN  INDUSTRIAL
           AND MUNICIPAL WASTEWATSRS
                       by
    J.S. Warner, T.M. Engel and P.J. Mondroa
         BatCelie Columbus Laboratories
              Columbus, Ohio  43201
            Contract No.  68-03-2956
                Project Officer
                Thomas Press ley
      Physical and Chemical Methods  Branch
Environmental Monitoring  and  Support Laboratory
             Cincinnati,  Ohio  45268
ENVIRONMENTAL MONITORING AND SUPPORT LABORATORY
       OFFICE OF RESEARCH AND DEVELOPMENT
      U.S. ENVIRONMENTAL PROTECTION AGENCY
             CINCINNATI, OHIO 45268

-------
                                DISCLAIMER
     This report has been reviewed by the Environmental Monitoring  and
Support Laboratory, U.S. Environmental Protection Agency, and approved
for publication.  Approval does not signify that the contents necessarily
reflect the view and policies of the U.S. Environmental Protection
Agency, nor does mention of trade names or commercial products constitute
endorsement or recommendation for use.
                                     ii

-------
                                   FOREWORD
    Environmental measurements are required to determine the quality of
ambient waters and the character of waste effluents.  The Environmental
Monitoring and Support Laboratory - Cincinnati, conducts research to:

    o    Develop and evaluate methods to measure the presence and concentra-
         tion of physical, chemical, and radiological pollutants in water,
         wastewater, bottom sediments, and solid wastes.

    o    Investigate methods for the concentration, recovery, and identifi-
         cation of viruses, bacteria, and other microbiological organisms in
         water; and, to determine the responses of aquatic organisms to
         water quality.

    o    Develop and operate an Agency-wide quality assurance program to
         assure standardization and quality control of systems for monitor-
         ing water and wastewater.

    o    Develop and operate a computerized system for instrument automation
         leading to improved data collection, analysis, and quality control.

    This report is one of a series that investigates the analytical behavior
of selected pesticides and suggests a suitable test procedure for their
measurement in wastewater.  The method was modeled after existing EPA
methods being specific yet as simplified as possible.
                                      Robert L. Booth, Director
                                      Environmental Monitoring and Support
                                      Laboratory - Cincinnati
                                   iii

-------
                                 ABSTRACT
     A method was developed  for the determination  of  rotenone in
wastewaters.  The method development  program  consisted  of a. literature
review; determination of extraction efficiency  for each compound from
water into me thylane chloride; development of a deactivated silica gel
cleanup procedure; and determination  of  a suitable high performance liquid
chromatographic (HPLC) method with ultraviolet  (UV) detection.

     The final method was applied to  a .relevant industrial wastewater to
determine the precision and  accuracy  of  the method.  The wastewater was
spiked with rotenone at levels of 5.5 pg/L and  110 yg/L.   Recovery for
rotenone at the 5.5 ug/L level was 85 ±  8 percent. Recovery at  the
110 ug/L level was 88 ± 3 percent.  The  method  detection limit  (MDL) for
rotenone in distilled water  was 1.6 ug/L.  The  MDL in wastewaters may be
higher due to interfering compounds.

     This report was submitted in partial fulfillment of Contract No.
68-03-2956 by Battelle Columbus Laboratories  under the  sponsorship of the
U.S. Environmental Protection Agency.  This report covers the period from
February 1, 1982 to April 30, 1982.
                                iv

-------
                                 CONTENTS

Foreword	
Abstract	        iv
Figures	        vi

     1.  Introduction	         1
     2.  Conclusions. 	         2
              Extraction and Concentration	         2
              Cleanup	'	         2
              Chromatography	         2
              Validation Studies	-	         2
     3.  Experimental 	         3
              Extraction and Concentration	         3
              Cleanup 	         3
              Chromatography	         3
              Validation Studies	         3
     4.  Results and Discussion 	         5
              Extraction and Concentration	         5
              Cleanup 	         5
              Chromatography	         5
              Validation Studies	         8

References	•      10
Appendix

       Rotenone  Method 6XX	        11

-------
                                 FIGURES

Number                                                          Page

  1    HPLC-UV Chromacogram of Standard Solution Representing
       5 yg/L of Rotenone in Water (Column 1).........     6

  2    HPLC-UV Chromatogram of Standard Solution Representing
       200 yg/L of Rotenone in Water (Column 2)	     7

  3    Analytical Curve for Rotenone	     9
                                   vi

-------
                                 SECTION 1

                               INTRODUCTION
     Rotenone (I) is a naturally occurring insecticide derived  from  derris
root.  It is used primarily to control insects on  food crops, but  is  also
used to eradicate undesirable fish species from  lakes and  streams.
                              OCH
 The CAS registry number for rotenone  is  £3-79-4  and  its  IUPAC  name  is
 [2R(-2a, 6a a, 12aa,)]l,2,12,12a-tetrahydro-8,9-dimethoxy-2-
 (1-methylethenyl) [l] benzopyrano(3,4-V)  furo[2,3-h]  [l]
 benzopyran-6(6a!O-one.  It has a melting  point of  165-166°C  and  an  oral
 LD50 in rats of 133 mg/Kg.  Common  synonyms  for  rotenone include "Derris"
 and "cube".  Several papers have been published  which  describe the
 analysis of rotenone using HPLC (1-6).  Most  of  the methods  use  reverse
 phase liquid chromatography with UV absorbance detection.  A method using
 normal phase HPLC has also been reported  (1).  The detection wavelengths
 include 254 nm (4), 280 nm (2-5), and 294 nm (1,3,4,5).   Methods for trace
 analysis using HPLC have been published  for  rotenone residues  on crops
 (4,5,6), in animal feed and tissues (3).  Cleanup  procedures using  silica
 gel have also been described (3,4,5,6).

      Rotenone is stable in water at neutral  pH and can be extracted from
 water with methylene chloride.  Rotenone  decomposes upon exposure to light
 and air; precautions should be taken  to  avoid excessive  exposure of
 rotenone containing solutions to light and air (7).  The selected approach
 to the determination of rotenone in water include  separatory funnel
 extraction from water with methylene chloride; cleanup using silica gel
 chromatography; and analysis using  HPLC with  UV  detection.   Standard
 concentration techniques using Kuderna-Danish (K-D) equipment  were  used.
 The final method is included in Appendix  A of this report.

-------
                                  SECTION  2

                                CONCLUSIONS
EXTRACTION AND CONCENTRATION

     Rotenone can be extracted  from water  into tnethylene  chloride with
greater than 90 percent recovery using  separatory funnel  techniques.  Use
of K-D concentration equipment  to  perform  extract concentrations  did  not
significantly affect compound recoveries.

CLEANUP

     Rotenone elutes from  six percent deactivated silica  gel  with greater
than 90 percent recovery.  This was an  effective  cleanup  procedure for a
relevant wastewater sample.

CHROMATOGRAPHY

     Two HFLC columns, Oupont Zorbax-Cyano (normal phase) and
Spherisorb-ODS (reversed phase), were found  to be acceptable  for  this
application.  The Dupont Zorbax-Cyano does not require  an additional
solvent exchange step and  was used as the  primary column.  The
Spherisorb-ODS column was  designated as  the  alternate column.

VALIDATION STUDIES

     Recoveries of rotenone from distilled water  in the 5.5 to  1090
Ug/L concentration range averaged  greater  than 85 percent.  The analytical
curve constructed from this data was linear.  The MDL in  distilled water
was 1.6 ug/L.  Recoveries  of rotenone from a pesticide,manufacturing
wastewater at the 5.5 and  110 ug/L levels  were 85 ± 8 and 88  ±  3  percent,
respectively.

-------
                                 SECTION 3

                               EXPERIMENTAL
     Studies were performed to determine if extractions with  separator?
funnels, cleanup by silica gel adsorption chromatography,  concentration
using K-D equipment, and analysis using HPLC with  UV detection  would  be
applicable technique for the determination of rotenone in  water.

EXTRACTION AND CONCENTRATION

     Extraction of rotenone from water was studied by using separatory
funnel techniques.  The distilled water was spiked with rotenone  at the
10 ug/L and 100 yg/L levels.  The sample was adjusted to pH 7 by  addition
of 1 ^ sodium hydroxide or 1 JJ sulfuric acid and extracted three  times
with 60 mL each of methylene chloride.  These studies were done in
duplicate.  The extracts were dried by passing  them through 10  cm of
anhydrous granular sodium sulfate, concentrated to five mL and  analyzed by
HPLC.  Column 1 was used for HPLC analyses.

CLEANUP

     A 10-gram silica gel column (six percent deactivaated with water) was
prepared as follows:  10 g silica gel was slurried with 50 mL of  acetone
containing 600 uL of reagent water, the slurry  was transferred  to a
chromatographic column and the solvent was allowed to elute and was
discarded.  The column was rinsed with 100 mL of methylene chloride which
was also discarded.  The rotenone, 10 or 100 us dissolved  in  one  mL of
methylene chloride, was added to the top of the column.  The  column was
eluted with 50 mL portions of methylene chloride (Fl), 6 percent  acetone
in methylene chloride (F2), 15 percent acetone  in  methylene chloride  (F3),
and 25 percent acetone in methylene chloride (F4). Each fraction was
concentrated to five mL and analyzed by HPLC.   Column 1 was used  for  HPLC
analyses.

CHROMATOGRAPHY

     Two HPLC columns were evaluated for the determination of rotenone:
Dupont Zorbax-Cyano and Spheriorb-ODS.

VALIDATION STUDIES

     The MDL for rotenone was determined by analyzing seven replicate
distilled water samples spiked at the 5.5 ug/L  concentration  levels  (8).
The sample extracts were cleaned up using the silica gel cleanup  procedure

-------
prior to analysis.  The amounts recovered were  determined  by external
standard calibration and the MDLs were calculated  from  these data.

     Distilled water was also spiked in duplicate  at  the 5.5,  22,  54,
270, and 1090 ug/L concentration levels and recoveries  of  the  rotenone
were determined as described earlier.  An analytical  curve was  generated
by plotting the amount spiked into the samples  versus the  amount recovered
from the samples.

     A relevant wastewater was used for wastewater validation  studies.
Seven replicates of the wastewater were analyzed to determine  the
background levels.  The wastewater was spiked with rotenone at  the  5.5  and
110 ug/L concentration levels, processed and analyzed.  Seven  replicate
extractions were performed at each concentration level.

-------
                                  SECTION  4

                          RESULTS  AND  DISCUSSION
EXTRACTION AND CONCENTRATION

     RoCenone was extracted from water  with  greater  than 85  percent
recovery using separatory  funnel techniques.   Recoveries of  rotentone from
water' using separatoy  funnels were  87 and  97  percent at  the  10 ug/L level
and 92 and 98 percent  at the  100 ug/L level.   These  data are results  of
duplicate analyses.

CLEANUP

     Rotenone eluted from  six percent deactivated  silica gel in the six
percent acetone in methylene  chloride fraction.  Recoveries  of 10 ug  and
100 yg of rotenone were 98 and  93 percent, respectively.

CHROMATOGRAPHY

     Both the Dupont Zorbax-Cyano and Spherisorb-QDS columns were
satisfactory for the determination  of rotenone.  Use of  the  Dupont
Zorbax-Cyano column requires  one less solvent  exchange  step  in the sample
workup and was chosen  as the  primary column.   The  following  conditions
were used:
     Column:

     Solvent:

     Flow:
     Detector:
     Injector Volume:
     Column:
     Solvent:
     Flow:
     Detector:
     Injector Volume:
Column 1

Dupont Zorbax-Cyano, 5 micron,
250 x 4.6 mm
30 percent methylene chloride,
70 percent hexane.
1 mL/min.
UV (§254 nm
10 uL

Column 2

spherisorb-ODS, 5 micron, 250 x 4.6 mm
60 percent acetonitrile, 40 percent water.
1 mL/min.
UV <§254 nm
10 UL
Chromatograms obtained under  these  conditions  are  shown  in Figures  1 and
2.

-------
 -|—i—r-'t—r f-r-i~-
 1.2      2.4      3.0
—|—i—i—i—r—|-T—r—r—«-

 4.0      6.0
 I •
7.2
                              Retention Time, min.
                                                      Rotenone
                                                  _A
-|—i—r- r--r -f—?•-
8.4     9.6
10.8    12.0
Figure 1.  HPLC-UV Chromatogram of Standard Solution Representing  5  ug/L
          of Rotenone  in Water (Column 1)

-------
                                                  Rotenone
                                                     L
•r~t"~t
1.5
3.0
4.5
-T-J •
  6.0
  I- -T— r-r-T'i~i
12.0      13.5
                                                7.5       9.0      10.5
                                            Retention Time,  mln.
          Figure  2.   HPLC-UV  Chromatogram of Standard Solution Representing  200 Mg/L
                     of  Uotenone  in Water (Column 2).
15.0

-------
VALIDATION STUDIES

     Recovery of rotenone from distilled water at  the 5.5 ug/L  level
was 4.6 ± 0.5 Ug/L.  The MDL in distilled water was calculated  to  be
1.6 Ug/L.  Recoveries of rotenone from distilled water at the 5.5,  22,
54, 270, and 1090 ug/L levels were 4.6(17), 20(20), 45(4.4),  240(10), and
950(2.3) Ug/L, respectively.  These data were the  averages of duplicate
analyses.  The percent relative range is given in.  parentheses.  The
resultant analytical curve is shown in Figure 3.
                                                   *
     Recoveries of rotenone from a relevant wasCawater at the 5.5  and
110 ug/L levels were 85 ± 8 percent and" 88 ± 3 percent, respectively.
These data were the averages of seven replicate analyses.  Rotenone was
not detected in the relevant wastewater.

-------
                1250  r
               1125  •
               1000  -
                875
Amount  Recovered,
     Ug/L
                750
                325   •
                500  •
                375
                250
                T25
                          I  .  I   .  I  .   I  .   I   .  I  .   I  .   I   .  I  .   I

                        125   250    375   500   625   750   875  1000  1125  1250


                                       Amount Spiked,  ug/L

                        Figure 3.  Analytical Curve  for Rotenone

-------
                                REFERENCES
                                f
1.  Bushway, R.J., B.D., Engdahi, B.M. Colvin, and A.R. Hanks.
    Separation of Rotenoids and the Determination of Rotenone
    in Pesticide Formulations by High Performance Liquid
    Chromatography. J. Assoc. Off. Anal. Chem., 58:965-970,
    1975.
                                                     t,
2.  Bushway, R.J. and A.R. Hanks, Determination of Rotenone
    in Pesticide Formulations and the Separation of Six Rotenoids
    by Reversed-phase High-Performance Liquid Chromatography.
    J. Chromatogr.  134:210-215, 1977. •

3.  Bowman, M.C., C.L., Holder, and L.I. Bone, High Pressure
    Liquid Chromatographic Determination of Rotenone and Degradation
    Products in Animal Chow and Tissues.  J. Assoc. Off. Anal.
    Chem. 16:1445-1455, 1978.

4.  Moring, I.E. and J.D. McHesney.  High Pressure Liquid
    Chromatographic Separation of Rotenoids from Plant Extracts.
    J. Assoc. Off. Anal. Chem. 62:774-781, 1979.

5.  Kobayeshi, H., 0. Matano, and G. Shinko.  Determination of
    Rotenoids in Soil and Crops by High-Performance Liquid
    Chromatography, J. Pesticide Sci. 5:89-92, 1980.

6.  Newsome, W.H. and J.B. Shields.  Residues of Rotenone and
    Rotenolone on Lettuce and Tomato Fruit after Treatment in the
    Field with Roteixone Formulations.  J. Agric. 'Food. Chem.
    28:722-724, 1980.

7.  Roteaone In: The Pesticide Manual, C.R. Worthing (ed.),
    BCPC Publications, Croydon, England, 1979.  p 468.

8.  Glaser, J.A. et. al. Trace Analysis for Wastewater.  Environmenta1
    Science and Technology 15:1426, 1981.
                                 10

-------
            DETERMINATION OF  ROTENONE  IN  INDUSTRIAL AND MUNICIPAL
                     WASTEWATERS BY LIQUID CHROMATOGRAPHY
                                  METHOD  635   ~	
1.   Scope and Application
     1.1   This method covers the determination of rotenone pesticide.  The
           following parameter can be determined by this method:
                     Parameters                          CAS No.
                     Rotenone                            83-79-4
     1.2   This is a high performance liquid chromatographic (HPLC) method
           applicable to the determination of'the compound listed above in
           municipal and industrial discharges as provided under  40 CFR 136.1.
           Any modification of this method beyond those expressly permitted
           shall be considered a major modification subject to application and
           approval of alternate test procedures under 40 CFR 136.4 and 136.5.
     1.3   The method detection limit (MDL, defined in Section 14)
           for rotenone compound is listed in Table 1.  The MDL for a specific
           wastewater may differ from those listed, depending upon the nature
           of interferences in the sample matrix.
     1.4   This method is restricted to use by or under the supervision of
           analysts experienced in the use of liquid chromatography and in the
           interpretation of liquid chromatograms.  Each analyst  must
           demonstrate the ability to generate acceptable results with this
           or muffle furnace at 400°C for 15-30 min.   Some thermally
                                       11

-------
     1.5   When this method is used to analyze unfamiliar samples for the
           compound above, compound identifications should be supported by at
           least one additional qualitative technique.  This method describes
           analytical conditions for a second liquid chromatographic column
           that can be used to confirm measurements made with the primary
           column.
2.   Summary of Method
     2.1   A measured volume of sample, approximately 1 liter, is solvent
           extracted with methylene chloride using a separatory funnel.
           Liquid chromatographic conditions are described which permit the
           separation and measurement of the compounds in the extract by HPLC-
           UV.l
3.   Interferences
     3.1   Method interferences may be caused by contaminants in solvents,
           reagents, glassware, and other sample processing hardware that lead
           to discrete artifacts or elevated baselines in liquid
           chromatograms.  All of these materials must be routinely
           demonstrated to be free from interferences under the conditions of
           the analysis by running laboratory reagent blanks as described in
           Section 8.5.
           3.1.1   Glassware must be scrupulously cleaned.2  Clean all
                   glassware as soon as possible after use by thoroughly
                   rinsing with the last solvent used in it.  Follow by
                   washing with hot water and detergent and thoroughly rinsing
                   with tap and reagent water.  Drain dry, and heat in an oven
                   or muffle furnace at 400°C for 15-30 min.  Some thermally
                                          12

-------
                   stable materials such as PCBs may not be eliminated by this
                   treatment.  Thorough rinsing with acetone and pesticide
                   quality hexane may be substituted for the heating.  After
                   drying and cooling, seal and store glassware in a clean
                   environment to prevent accumulation of dust or other
                   contaminants.  Store inverted or capped with aluminum foil.
           3.1.2   The use of high purity reagents and solvents helps to
                   minimize interference problems.  Purification of solvents by
                   distillation in all-glass systems may be required.
     3.2   Matrix interferences may be caused by contaminants that are
           coextracted from the sample.  The extent of matrix interferences
           will vary considerably from source to source, depending upon the
           nature and diversity of the industrial complex or municipality being
           sampled.  The cleanup procedure in Section 11 can be used to
           overcome these interferences, but unique samples may require
           additional cleanup approaches to achieve the MDL listed in Table 1.
4.   Safety
     4.1   The toxicity or carcinogenicity of each reagent used in this method
           has not been precisely defined; however, each chemical compound
           should be treated as a potential health hazard.  From this
           viewpoint, exposure to these chemicals must be reduced to the
           lowest possible level by whatever means available.  The laboratory
           is responsible for maintaining a current awareness file of OSHA
           regulations regarding the safe handling of the chemicals specified
           in this method.  A reference file of material data handling sheets
           should also be made available to all personnel involved in the
                                        13

-------
           chemical analysis.  Additional references to laboratory safety are
           available and have been identified3^ for the information of the
           analyst.
5.   Apparatus and Materials
     5.1   Sampling equipment, for discrete or composite sampling.
           5.1.1   Grab sample bottle - Amber borosilicate or flint glass,
                   1-liter or 1-quart volume, fitted with screw caps lined
                   with Teflon.  Aluminum foil may be substituted for Teflon
                   if the sample is not corrosive.  If amber bottles are not
                   available, protect sample's from light.  The container and
                   cap liner must be washed, rinsed with acetone or methylene
                   chloride, and dried before use to minimize contamination.
           5.1.2   Automatic sampler (optional) - Must incorporte glass sample
                   containers for the collection of a minimum of 250 ml.
                   Sample containers must be kept refrigerated at 4°C and
                   protected from light during compositing.  If the sampler
                   uses a peristaltic pump, a minimum length of compressible
                   silicons rubber tubing may be used.  Before use, however,
                   the compressible tubing should be thoroughly rinsed with
                   methane!, followed by repeated rinsings with distilled water
                   to minimize the potential for contamination of the sample.
                   An integrating flow meter is required to collect flow
                   proportional composites.
     5.2   Glassware (All specifications are suggested.  Catalog numbers are
           included for illustration only.)
           5.2.1   Separatory funnel - 2000-mL, with Teflon stopcock.
                                         14

-------
      5.2.2   Drying column - Chromatographic column 400 mm long x 10 mm
              ID with coarse frit.
      5.2.3   Chromatographic column - 400 mm long x 19 mm ID with 250 ml
              reservoir at the top and Teflon stopcock (Kontes K-420290)
              or equivalent).
      5.2.4   Concentrator tube, Kuderna-Danish - 25-mL, graduated
              (Kontes K-570050-2525 or equivalent).  Calibration must be
              checked at the volumes employed in the test.  A ground glass
              stopper is used to prevent evaporation of extracts.
      5.2.5   Evaporative flask, Kuderna-Danish - 250-mL (Kontes
              K-570001-0250 or equivalent).  Attach to concentrator tube
              with springs.
      5.2.6   Snyder column, Kuderna-Danish - three-ball macro (Kontes
              K-503000-0121 or equivalent).
      5.2.7   Snyder column, Kuderna-Danish - two-ball micro (Kontes
              K-569001-0219 or equivalent).
      5.2.8   Vials - Amber glass, 5 to 10 ml capacity with Teflon lined
              screw-cap.
      5.2.9   Volumetric flask - 10 mL.
      5.2.10  Erlenmeyer flask - 250 mL.
      5.2.11  Graduated cylinder - 1000 mL.
5.3   Boiling chips - approximately 10/40 mesh carborundum.  Heat to
      400°C for 4 hours or extract in a Soxhlet extractor with methylene
      chloride.
5.4   Water bath - Heated, capable of temperature control (^2°C).  The
      bath should be used in a hood.
                                     15

-------
     5.5   Balance - Analytical, capable of accurately weighing to the nearest
           0.0001 g.
 -   5.6   Liquid chromatograph - Analytical system complete with liquid
           chromatograph and all required accessories including syringes,
           analytical columns, detector and strip-chart recorder.  A data
           system is recommended for"measuring peak areas.
           5.6.1   Pump - Isocratic pumping system, constant flow.
           5.6.2   Column 1 - Normal-phase column, 5 micron Zorbax-CN,
                   250 x 4.6 mm or equivalent.
           5.6.3   Column 2 - Reversed-phase column, 5 micron Spherisorb-ODS,
                   250 x 4.6 mm or equivalent.
           5.6.4   Detector - Ultraviolet absorbance detector, 254 nm.
6.   Reagents
     6.1   Reagent water - Reagent water is defined as a water in which an
           interferent is not observed at the method detection limit of each
           parameter of interest.
     6.2   Methylene chloride, methanol, acetonitrile, acetone, hexane,
           distilled-in-glass quality or equivalent.
     6.3   Sodium sulfate (ACS) Granular, anhydrous; heated in a muffle
           furnace at 400°C overnight.
     6.4   IN. sulfuric acid.
     6.5   IN sodium hydroxide.
     6.6   Silica gel, Davison grade 923, 100-200 mesh, dried for 12 hours at
           150°C.
     6.7   Stock standard solutions (l.OOug/L) - Stock standard solutions can
           be prepared from pure standard materials or purchased as certified
           solutions.
                                        16

-------
           6.7.1   Prepare stock standard solutions by accurately weighing
                   about 0.0100 grams of pure material.  Dissolve the material
                   in distilled-in-glass quality methylene chloride for
                   analyses performed using column 1 and methanol for analyses
                   performed using column 2.  Dilute to volume in a 10-mL
                   volumetric flask.   Larger volumes can be used at the
                   convenience of the analyst.  If compound purity is certified
                   at 96% or greater, the weight can be used without
                   correction to calculate the concentration of the stock
                   standard.  Commercially prepared stock standards can be used
                   at any concentration if they are certified by the
                   manufacturer or by an independent source.
           6.7.2   Transfer the stock standard solutions into Teflon-sealed
                   screw-cap bottles.  Store at 4°C and protect from light.
                   Stock standard solutions should be checked frequently for
                   signs of degradation or evaporation, especially just prior
                   to preparing calibration standards from them.
           6.7.3   Stock standard solutions must be replaced after six months
                   or sooner if comparison with check standards indicates a
                   problem.
7.   Calibration
     7.1   Establish liquid chromatographic operating parameters equivalent to
           those indicated in Table 1.  The liquid chromatographic system can
           be calibrated using the external standard technique (Section 7.2)
           or the internal standard technique (Section 7.3).
                                        17

-------
7.2   External standard calibration procedure:
      7.2.1   For each compound of interest, prepare calibration
              standards at a minimum of three concentration levels by
              adding volumes of one or more stock standards to a
              volumetric flask and diluting to volume with 50/50
              hexane/methylene chloride for column 1 standards and
              acetonitrile for column' 2 standards.  One of the external
              standards should be at a concentration near, but above, the
              method detection limit.  The other concentrations should
              correspond to the expected range of concentrations found in
              real samples or should define the working range of the
              detector.
      7.2.2   Using injections of 5 to 20ul_ of each calibration
              standard, tabulate peak height or area responses against
              the mass injected.  The results can be used to prepare a
              calibration curve for each compound.  Alternatively, the
              ratio of the response to the mass injected, defined as the
              calibration factor (CF), can be calculated for each compound
              at each standard concentration.  If the relative standard
              deviation of the calibration factor is less than 10% over
              the working range, linearity through the origin can be
              assumed and the average calibration factor can be used in
              place of a calibration curve.
      7.2.3   The working calibration curve or calibration factor must be
              verified on each working shift by the measurement of one or
              more calibration standards.  If the response for any

                                      18

-------
              compound varies from the predicted response by more than
              ^10%, the test must be repeated using a fresh calibration
              standard. "Alternatively, a new calibration curve or
              calibration factor must be prepared for that compound.
7.3   Internal standard calibration procedure.  To use this approach, the
      analyst must select one or more internal standards similar in
      analytical behavior to the compounds of interest.  The analyst must
      further demonstrate that the measurement of the internal standard is
      not affected by method or matrix interferences.  Due to these
      limitations, no internal standard applicable to all samples can be
      suggested.
      7.3.1   Prepare calibration standards at a minimum of three
              concentration levels for each compound of interest by
              adding volumes of one or more stock standards to a
              volumetric flask.  To each calibration standard, add a known
              constant amount of one or more internal standards, and
              dilute to volume with 50/50 hexane/methylene chloride for
              column 1 standards and acetonitrile for column 2 standards.
              One of the standards should be at a concentration near, but
              above, the method detection limit.  The other
              concentrations should correspond to the expected range of
              concentrations found in real samples, or should define the
              working range of the detector.
      7.3.2   Using injections of 5 to 2QuL of each calibration
              standard, tabulate the peak height or area responses against
              the concentration for each compound and internal standard.
                                     19

-------
              Calculate response factors (RF) for each compound as
              follows:
                            RF • (AsCis)/(AisCs)
              where:
                   As  » Response for the compound to be measured.
                   A-jS * Response for the internal standard.
                   Cis = Concentration of the internal standard inug/L.
                   Cs  = Concentration of the compound to be measured in
                         ug/L.
              If the RF value over the working range is constant, less
              than 10% relative standard deviation, the RF can be assumed
              to be invariant and the average RF can be used for
              calculations.  Alternatively, the results can be used to
              pilot a calibration curve of response ratios, As/Ais
              against RF.
      7.3.3   The working calibration curve or RF must be verified on
              each working shift by the measurement of one or more
              calibration standards.  If the response for any compound
              varies from the predicted response by more than ^10%, the
              test must be repeated using a fresh calibration standard.
              Alternatively, a new calibration curve must be prepared for
              that compound.
7.4-   Before using any cleanup procedure, the analyst must process a
      series of calibration standards through the procedure to validate
      elution patterns and the absence of interferences from the
      reagents.
                                     20

-------
8.   Quality control
     8.1   Each laboratory using this method is required to operate a formal
           quality control program.  The minimum requirements of this program
           consist of an initial demonstration of laboratory capability and the
           analysis of spiked samples as a continuing check on performance.
           The laboratory is required to maintain performance records to define
           the quality of data that is generated.
           8.1.1   Before performing any analyses, the analyst must
                   demonstrate the ability to generate acceptable accuracy and
                   precision with this method.  This ability is established as
                   described in Section 8.2.
           8.1.2   In recognition of the rapid advances occurring in
                   chromatography, the analyst is permitted certain options to
                   improve the separations or lower the cost of measurements.
                   Each time such modifications to the method are made, the
                   analyst is required to repeat the procedure in Section 8.2.
           8.1.3   The laboratory must spike and analyze a minimum of 10% of
                   all samples to monitor continuing laboratory performance.
                   This procedure is described in Section 8.4.
     8.2   To establish the ability to generate acceptable accuracy and
           precision, the analyst must perform the following operations.
           8.2.1   Select a representative spike concentration for each
                   compound to be measured.  Using stock standards, prepare a
                   quality control check sample concentrate in methylene
                   chloride 1000 times more concentrated than the selected
                   concentrations.
                                           21

-------
      8.2.2   Using a pipet, add 1.00 ml of the check sample concentrate
              to each of a minimum of four 1000-ml aliquots of reagent
              water.  A representative wastewater may *e used in place of
              the reagent water, but one or more additional aliquots must
              be analyzed to determine background levels, and the spike
              level must exceed twice the background level for the test
              to be valid.  Analyze the aliquots according to the method
              beginning in Section 10.
      8.2.3   Calculate the average percent recovery (R), and the
              standard deviation of the percent recovery (s), for the
              results.  Wastewater background corrections must be made
              before R and s calculations are performed.
      8.2.4   Using the appropriate data from Table 2, determine the
              recovery and single operator precision expected for the
              method, and compare these results to the values measured in
              Section 8.2.3.  If the data are not comparable, the analyst
              must review potential problem areas and repeat the test.
8.3   The analyst must calculate method performance criteria and define
      the performance of the laboratory for each spike concentration and
      parameter being measured.
      8.3.1   Calculate upper and lower control limits for method
              performance as follows:
                       Upper Control Limit (UCL) = R +• 3 s
                       Lower Control Limit (LCL) = R - 3 s
              where R and s are calculated as in Section 8.2.3.  The UCL
              and LCL can be used to construct control charts** that are
              useful in observing trends in performance.
                                      22

-------
      8.3.2   The laboratory must develop and maintain separate accuracy
              statements of laboratory performance for wastewater
              samples.  An accuracy statement for the method is defined
              as R £ s.  The accuracy statement should be developed by the
              analysis of four aliquots of wastewater as described in
              Section 8.2.2, followed by the calculation of R and s.
              Alternately, the analyst may use four wastewater data points
              gathered through the requirement for continuing quality
              control in Section 8.4.  The accuracy statements should be
              updated with this method.  This ability is established as
              described regularly.5
8.4   The laboratory is required to collect in duplicate a portion of
     »their samples to monitor spike recoveries.  The frequency of spiked
      sample analysis must be at least 10% of all samples or one sample
      per month, whichever is greater.  One aliquot of the sample must be
      spiked and analyzed as described in Section 8.2.  If the recovery
      for a particular compound does not fall within the control limits
      for method perfomance, the results reported for that compound in
      all samples processed as part of the same set must be qualified as
      described in Section 13.3.  The laboratory should monitor the
      frequency of data so qualified to ensure that it remains at or
      below 5%.
8.5   Before processing any samples, the analyst should demonstrate
      though the analysis of a 1-liter aliquot of reagent water that all
      glassware and reagents interferences are under control.  Each time
                                       23

-------
           a set of samples is extracted or there is a change in reagents, a
           laboratory reagent blank should be processed as a safeguard against
           laboratory contamination.
     8.6   It is recommended that the laboratory adopt additional quality
           assurance practices for use with this method.  The specific
           practices that are most productive depend upon the needs of the
           laboratory and the nature of the samples.  Field duplicates may be
           analyzed to monitor the precision of the sampling technique.  When
           doubt exists over the identification of a peak on the chromatogram,
           confirmatory techniques such as liquid chromatography with a
           dissimilar column, must be used.  Whenever possible, the laboratory,
           should perform analysis of standard reference materials and
           participate in relevant performance evaluation studies.
9.   Samples Collection, Preservation, and Handling
     9.1   Grab samples must be collected in glass containers.  Conventional
           sampling practices? should be followed; however, the bottle must
           not be prerinsed with sample before collection.  Composite samples
           should be collected in refrigerated glass containers in accordance
           with the requirements of the program.  Automatic sampling equipment
           must be as free as possible of Tygon and other potential sources of
           contamination.
     9.2   The samples must be iced or refrigerated at 4°C from the time of
           collection until extraction.
     9.3   Adjust the pH of the sample to 6 to 8 with 1N_ sodium hydroxide or
           IN sulfuric acid immediately after sampling.
                                          24

-------
10.  Sample Extraction
     10.1  Mark the water meniscus on the side of the sample bottle for later
           determination of sample volume.  Pour the entire sample into a
           2-liter separatory funnel.  Check the pH of the sample with wide
           range pH paper and adjust to 7 with 1 N sodium hydroxide or 1 N
           H2S04-
     10.2  Add 60 ml of methylene chloride-to the sample bottle,  seal, and
           shake 30 seconds to rinse the inner walls.  Transfer the solvent to
           the separatory funnel  and extract the sample by shaking the funnel
           for 2 minutes with periodic venting to release excess  pressure.
           Allow the organic layer to separate from the water phase for a
           minimum of 10 minutes.  If the emulsion interface between layers is
           more than one third the volume of the solvent layer, the analyst
           must employ mechanical techniques to complete the phase separation.
           The optimum technique  depends upon the sample, but may include
           stirring, filtration of the emulsion through glass wool,
           centrifugation,  or other physical methods.  Collect the methylne
           chloride extract in a  250-mL Erlenmeyer flask.
     10.3  Add a second 60-mL volume of methylene chloride to the sample
           bottle and repeat the  extraction  procedure a second time,
           collecting the extract.   Perform  a third extraction in the same
           manner and collect the extract.
     10.4  Assemble a Kuderna-Danish (K-0) concentrator by attaching a 25-mL
           concentrator tube to a 250-mL evaporative flask.   Other
           concentration devices  or techniques may be used in place of the K-D
           if  the requirements  of Section 8.2 are met.
                                          25

-------
10.5  Pour the combined extract through a drying column containing about
      10 cm of anhydrous sodium sulfate, and collect the extract in the
      K-D concentrator.  Rinse the Erlenmeyer flask and column with 20 to
      30 ml of methylene chloride to complete the quantitative transfer.
      Once the flask rinse has passed through the drying column, rinse
      the column with an additional 30 to 40 mL of methylene chloride.
10.6  Add 1 or 2 clean boiling chips to the evaporative flask and attach
      a three-ball Snyder column.  Prewet the Snyder column by adding
      about 1 mL methylene chloride to the top.  Place the K-D apparatus
      on a hot water bath, 60 to 65°C, so that the concentrator tube is
      partially immersed in the hot water, and the entire lower rounded
      surface of the flask is bathed with hot vapor.  Adjust the vertical
      position of the apparatus and «the water temperature as required to
      complete the concentration in 15 to 20 minutes.  At the proper rate
      of distillation, the balls of the column will actively chatter but
      the chambers will not flood with condensed solvent.  When the
      apparent volume of liquid reaches 1 mL, remove the K-D apparatus
      and allow it to drain and cool for at least 10 minutes.
10.7  Remove the macro-Snyder column and rinse the flask and its lower
      joint into the concentrator tube with 1 to 2 mL of methylene
      chloride.  A 5-mL syringe is recommended for this operation.  Add 1
      or 2 clean boiling chips and attach a two-ball micro-Snyder column
      to the concentrator tube.  Prewet the micro-Snyder column with
      methylene chloride and concentrate the solvent extract as before.
      When an apparent volume of 0.5 mL is reached, or the solution stops
      boiling, remove the K-D apparatus and allow it to drain and cool

                                     26

-------
           for 10 minutes.   If analysis  is  being performed using column 1  or
           if sample cleanup is required,  proceed with Section  10.9.   If column  2
           is used and no sample cleanup is required,  proceed with Section 10.8.
     10.8  Add 10 ml of acetonitrile to  the concentrator tube along with 1 or
           2 clean boiling chips.  Attach a two-ball micro-Snyder column to
           the concentrator tube.  Prewet the micro-Snyder column with
           acetonitrile and concentrate  the solvent extract as  before.  When
           an apparent volume of 1 ml is reached, remove the K-D apparatus and
           allow it to drain and cool for 10 minutes.   Transfer the liquid to
           a 10-mL volumetric flask and  dilute to the mark with acetonitrile.
           Mix thoroughly prior to analysis.  Proceed with Section 12  using
           column 2.
     10.9  Remove the micro-Snyder column and adjust the volume of the extract
           to 1.0 ml with methylene chloride.  Stopper the concentrator tube
           and store refrigerated if further processing will not be performed
           immediately.  If the extract  is to be stored longer  than two days,
           transfer the extract to a screw-capped vial with a Teflon-lined
           cap.  If the sample extract requires no further cleanup, proceed
           with the liquid chromatographic analysis in Section  12 using Column 1.
           If the sample requires cleanup,  proceed to  Section 11.
     10.10 Determine the original sample volume by refilling the sample bottle
           to the mark and transferring  the water to a 1000-mL  graduated
           cylinder.  Record the sample  volume to the nearest 5 mL.
11.  Cleanup and Separation
     11.1  Cleanup procedures may not be necessary for a relatively clean
           sample matrix.  The cleanup procedure recommended in this method
           has been used for the analysis  of various clean waters and
                                          27

-------
      industrial effluents.  If particular circumstances demand the use
      of additional cleanup, the analyst must demonstrate that the
      recovery of each compound of interest is no less than 85%.
11.2  Slurry 10 g of silica gel in 50 ml of acetone to which has been
      added SOOuL of reagent water.  Transfer the slurry to a
      chromatographic column (silica gel is retained with a plug of glass
      wool).  Wash the column with 100 ml of methylene chloride.  Use a
      column flow rate of 2 to 2.5 mL/min throughout the wash and elution
      profiles.-
11.3  Add the extract from Section 10.9 to the head of the column.  Allow
      the solvent to elute from the column until the silica gel is almost
      exposed to the air.  Elute the column with 50 mL of methylene
      chloride.  Discard this fraction.
11.4  Elute the column with 60 ml of 6% acetone in methylene chloride.
      Collect this fraction in a K-D apparatus.  Concentrate the column
      fraction to 1 ml as described in Sections 10.6 and 10.7.  If column
      1 is being used, proceed with Section 11.5.  If column 2 is being
      used, proceed with Section 11.7
11.5  Add 5 mL of hexane to the concentrate along with 1 or 2 clean
      boiling chips.  Attach a three-ball micro-Snyder column to the
      concentrator tube.  Prewet the micro-Snyder column with hexane and
      concentrate the solvent extract to an apparent volumn of 1 mL.
      Allow the K-D apparatus to drain and cool for 10 minutes.
11.6  Transfer the liquid to a 10-mL volumetric flask and dilute to the
      mark with hexane.  Mix thoroughly prior to analysis.  If the
      extracts will not be analyzed immediately, they should be
                                   28

-------
           transferred to Teflon sealed screw-cap vials and refrigerated.
           Proceed with the liquid chromatographic analysis using column 1.
     11.7  Add 10 mL of acetonitrile to the concentrate along with 1 or 2
           clean boiling chips.  Attach a three-ball micro-Snyder column to
           the concentrator tube.  Prewet the micro-Snyder column with
           acetonitrile and concentrate the solvent extract to an apparent
           volume of 1 mL.  Allow the K-D apparatus to drain and cool for 10
           minutes.
     11.8  Transfer the liquid to a 10-mL volumetric flask and dilute to the
           mark with acetonitrile.  Mix thoroughly prior to analysis.  If the
           extracts will not be analyzed immediately, they should be
           transferred to Teflon sealed screw-cap vials and refrigerated.
           Proceed with the liquid chromatographic analysis using colunjn 2.
12.  Liquid Chromatography
     12.1  Table 1 summarizes the recommended operating conditions for the
           liquid chromatograph.  Included in this table are estimated
           retention times and method detection limits that can be achieved by
           this method.  An example of the separations achieved by column 1 and
           column 2 are shown in Figures 1 and 2.  Other columns,
           chromatographic conditions, or detectors may be used if the
           requirements of Section 8.2 are met.
     12.2  Calibrate the liquid chromatographic system daily as described in
           Section 7.
     12.3  If an  internal  standard approach is being used, the analyst must
           not add the internal  standard to sample extracts until immediately
           before injection into the instrument.   Mix thoroughly.
                                        29

-------
     12.4  Inject 5 to 20UL of the sample extract by completely filling the
           sample valve loop.  Record the resulting peak sizes in area or peak
           height units.
     12.5  The width of the retention time window used to make identifications
           should be based upon measurements of actual retention time
           variations40f standards over the course of a day.  Three times the
           standard deviation of a retention time for a compound can be used
           to calculate a suggested window size; however, the experience of
           the analyst should weigh heavily in the interpretation of
           chromatograms.
     12.6  If the response for the peak exceeds the working range of the
           system, dilute the extract and reanalyze.
     12.7  If the measurement of the peak response is prevented by the
           presence of interferences, further cleanup is required.
13.  Calculations
     13.1  Determine the concentration of individual compounds in the sample.
           13.1.1  If the external standard calibration procedure is used,
                   calculate the amount of material injected from the peak
                   response using the calibration curve or calibration factor
                   in Section 7.2.2.  The concentration in the sample can be
                   calculated as follows:
                                              (A)(Vt)
                        Concentration, ug/L a 	
                                              (V1)(VS)
                   where:
                     A    = Amount of material injected, in nanograms.
                     V-j    s Volume of extract injected inuL.
                                          30

-------
                     Vt s Volume of total  extract inuL.
                     Vs = Volume of water  extracted in ml.
           13.1.2  If the internal standard calibration procedure was used,
                   calculate the concentration in the sample using the
                   response factor (RF) determined in Section 7.3.2 as
                   follows:
                                                (As)(is)
                        Concentration, yg/L
s
                                              (Ais)(RF)(V0)
                   where:
                     As  = Response for the compound to be measured.
                     ATS - Response for the internal standard.
                     Is  - Amount of internal  standard added to each extract
                           extract i n u g.
                     V0  = Volume of water extracted, in liters.
     13.2  Report results in micrograms per liter without correction for
           recovery data.  When duplicate and  spiked samples are analyzed,
           report all data obtained with the sample results.
     13.3  For samples processed as part of a  set where the laboratory spiked
           sample recovery falls outside of the control limits in Section 8.3,
           data for the affected compounds must be labeled as suspect.
14.  Method Performance
     14.1  The method detection limit (MDL) is defined as the minimum
           concentration of a substance that can be measured and reported with
           99% confidence that the value is about zero.  The MDL
           concentrations listed in Table 1 were obtained using reagent
           water.1  Similar results were achieved using representative
           wastewaters.
                                           31

-------
14.2  This method has been tested for linearity of recovery from spiked
      reagent water and has been demonstrated to be applicable over the
      concentration range from 10 x MDL to 1000 x MOL.
14.3  In a single laboratory, Battelle Columbus-Laboratories, using
               (. j
      spiked waste.^ater samples, the average recoveries presented in
      Table 2 were obtained.  Seven replicates of each of two different
      wastewaters were spiked and analyzed.  The standard deviation of the
      percent recovery is also included in Table 2.1
                                     32

-------
U)
CJ
1.2
                                                                               Rotenone
                               2.4
 I  •  '
3.0
~r — | —|
  6.0
     4.8
Retention Time, min.
7.2
"1—r~r» f-i
 8.4
                                                                                             T—r
9.0     10.0    12.0
                    FIGURE 1.  IIPLC-UV CIIROHATOGRAM OF STANDARD SOLUTION REPRESENTING  5  M9/L
                               OF ROTENONE IN WATER (COLUMN 1).

-------
                                                            Rotenone
U>
                   1.5       3.0      4.5       6.0       7.5       9.0      10.5      12.0     13.5    15.0

                                            Retention Time,  min.

                   FIGURE 2.   IIPLC-UV CHROMATOGRAM OF STANDARD SOLUTION REPRESENTING 200 Mq/L OF
                              ROTENONE IN WATER (COLUMN 2).

-------
      TABLE 1.  CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
                       Retention Time (min.)       Method Detection Limit
  Parameter            Column 1Column 2-                (
Rotenone                  8.6        8.0                     1.6
Rotenone

Column 1 conditions:  Zorbax-CN, 5 micron, 250 x 4.6 mm;  1 mL/min flow; 30/70
methylene chloride/hexane.

Column 2 conditions:  Spherisorb-ODS, 5 micron, 250 x 4.6 mm;  1 mL/min flow;
60/40 acetonitrile/water.
                                           35

-------
         TABLE 2.  SINGLE LABORATORY ACCURACY AND PRECISION^)
Parameter
Average
Percent
Recovery
                            Standard
                            Deviation,
           Spike
           Level
         Number
           of
        Analyses
           Matrix
           Type(b)
Rotenone
  85
  88
8
3
5.5
109
7
7
(a)  Column 1 conditions were used.
(b)  1 » __pesticide^manufacturing  wastewater
                                         36

-------
                                 REFERENCES
1.  "Development of Methods for Pesticides in Wastewaters," Report for EPA
    Contract 68-03-2956 (In preparation).

2.  ASTM Annual Book of Standards, Part 31, D3694, "Standard Practice for
    Preparation of Sample Containers and for Preservation," American Society
    for Testing and Materials, Philadelphia, PA, p. 679, 1980.

3.  "Carcinogens - Working with Carcinogens," Department of Health, Education,
    and Welfare, Public Health Service, Center for Disease Control, National
    Institute for Occupational Safety and Health, Publication No. 77-206,
    August, 1977.

4.  "OSHA Safety and Health Standards, General  Industry," (29 CFR 1910),
    Occupational Safety and Health Adninistrati on, OSHA 2206 (Revised,
    January, 1976).

5.  "Safety in Academic Chemistry Laboratories," American Chemical Society
    Publication, Committee on Chemical Safety,  3rd Edition, 1979.

6.  "Handbook for Analytical  Quality Control  in Water and Wastewater
    Laboratories," EPA-600/4-79-019, U. S. Environmental Protection Agency,
    Environmental Monitoring  and Support Laboratory - Cincinnati, Ohio 45268,
    March, 1979.

7.  ASTM Annual Book of Standards, Part 31, D3370, "Standard Practice for
    Sampling Water," American Society for Testing and Materials, Philadelphia,
    PA, p. 76, 1980.

8.  Glaser, J. A. et al, "Trace Analysis for Wastewaters," Environmental
    Science and Technology. 1_5, 1426 (1981).
                                       37

-------
                                    TECHNICAL REPORT DATA
                            (Please read Instructions on the reverse before completing)
1. REPORT NO.
                              2.
                                                            3. RECIPIENT'S ACCESSION NO.
4. TITLE AND SUBTITLE
                                                            S. REPORT DATE
   Determination of Rotenone in Industrial  and
   Municipal Wastewaters
             6. PERFORMING ORGANIZATION CODE
7. AUTHOR(S)

  J.S.  Warner, T.M.  Engel, and P.O. Mondron
                                                            8. PERFORMING ORGANIZATION REPORT NO.
9. PERFORMING ORGANIZATION NAME AND ADDRESS
  Battene Columbus Laboratories
  505  King Avenue
  Columbus, Ohio  43201
             10. PROGRAM ELEMENT NO.

                CBEBIC
             11. CONTRACT/GRANT NO.
                                                               68-03-2956
12. SPONSORING AGENCY NAME AND ADDRESS
   Environmental Protection Agency
   Environmental Monitoring and Support  Laboratory
   Cincinnati, Ohio    45268
              13. TYPE OF REPORT AND PERIOD COVERED
                2/82 - 4/82            	
             14. SPONSORING AGENCY CODE


                EPA 600/6
IS. SUPPLEMENTARY NOTES
16. ABSTRACT
     A method was developed  for the determination of rotenone in wastewaters.   The
 method  development program  consisted of  a  literature review; determination  of
 extraction  efficiency for each compound  from water into metnylene chloride;
 development of a deactivated- silica gel  cleanup procedure;  and determination  of a
 suitable  high performance liquid chromatographic (HPLC) method with ultraviolet
 (UV) detection.

     The final method was  applied to a relevant industrial wastewater to determine
 the precision and accuracy  of the method.   The wastewater was spiked with rotenone
 at levels of 5.5 wg/L and 110 wg/L.  Recovery for rotenone  at the 5.5 wg/L  level
 was 85  *  8  percent.  Recovery at the 110 yg/L level was 88  * 3 percent.  The  method
 detection limit (MDL) for rotenone in distilled water was- 1.6'wg/L.  The MDL  in
 wastewaters may be higher due to interfering compounds.

                                                   U.S. Environmental Protection
                                                   Region 5, Library (PL-12J)
                                                   11 West Jackson Boulevard, 12th
                                                   Chicago, IL  60604-3590
17.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
                                               b.IDENTIFIERS/OPEN ENDED TERMS
                           c.  COSATI Field/Group
18. DISTRIBUTION STATEMENT
  Distribute to Public
19. SECURITY CLASS (Tills Report/
  Not Classified	
                                                                           21. NO. Or PAGES
                                                                                 37
20. SECURITY CLASS {Tliis page/

  Not Classified	
                                                                           22. PRICE
EPA Form 2220-1 (R«». 4-77)   PKCVIOUS  SOITION is OBSOLETE

-------