600485024                                  DRAFT
                 DETERMINATION  OF ORYZALIN IN INDUSTRIAL
                          AND MUNICIPAL WASTEWATERS
                                METHOD  638
1.   Scope and Application
    1.1  This method covers the determination  of oryzalin pesticide.  The
        following parameter can be determined  by this method:
                  Parameters                         CAS No.
                  Oryzalin                         19044-88-3
    1.2  This is a high  performance liquid chromatographic (HPLC) method
        applicable to the determination of the compound listed above in
        municipal and industrial discharges as provided under  40 CFR 136.1.
        Any modification of this method beyond those expressly permitted
        shall be considered a major modification subject to application and
        approval of alternate test procedures  under 40 CFR 136.4 and 136.5.
    1.3  The estimated method  detection limit  (MDL, defined in Section  15)
        for oryzalin is listed  in Table 1. The MDL for a specific waste-
        water may differ from those listed, depending  upon the nature  of
        interferences in the  sample matrix.

-------
    1.4  This method is restricted to use by or under the supervision of
         analysts experienced in the use of liquid  chromatography and -in
         the interpretation of liquid chromatograms.   Each analyst must
         demonstrate the ability to generate acceptable results  with  this
         method using the procedure described in Section 8.2.
    1.5  When this method is used to analyze unfamiliar samples  for the
         compound above, compound identifications should be supported by at
         least one additional qualitative technique.   This method describes
         analytical conditions for a second liquid  chromatographic column
         that can be used to confirm measurements made with the  primary
         column.
2.  Summary of Method
    2.1  A measured volume of sample, approximately 1 liter, is  solvent  ex-
         tracted with methylene chloride using  a separatory funnel.   The
         methylene chloride extract is dried and exchanged to  acetonitrile
         during concentration to a volume of 2  ml or  less.  Liquid chromato-
         graphic conditions are described which permit the separation and
         measurement of the compounds in the extract  by HPLC-UV.1
3.  Interferences
    3.1  Method  interferences may be caused by  contaminants in solvents,
         reagents, glassware, and other sample  processing hardware that  lead
         to discrete artifacts or elevated baselines  in liquid chromatograms.
         All of these materials must be routinely demonstrated to be  free
         from interferences under the conditions of the analysis  by running
         laboratory reagent blanks as described in  Section 8.5.

-------
         3.1.1  Glassware must be scrupulously cleaned2.  Clean all glass-
                ware as soon as possible after use by thoroughly rinsing with
                the last solvent used in it.  Follow by washing with hot water
                and detergent and thorough rinsing with tap and reagent water.
                Drain dry, and heat in an oven or muffle furnace at 400°C for
                15 to 30 min.  Some thermally stable materials such as PCBs may
                not be eliminated by this treatment.  Thorough rinsing with
                acetone and pesticide quality hexane may be substituted for
                muffle furnace heating.   Volumetric wave should not be heated
                in a muffle furnace.  After drying and cooling, seal  and store
                glassware in a clean environment to prevent any accumulation
                of dust or other contaminants.  Store inverted or capped with
                aluminum foil.
         3.1.2  The use of high purity reagents and solvents helps to mini-
                mize interference problems.  Purification of solvents by
                distillation in all-glass systems may be required.
    3.2  Matrix interferences may be caused by contaminants that are  coex-
         tracted from the sample.  The extent of matrix interferences will
         vary considerably from source to source, depending upon the  nature
         and diversity of the industrial  complex or municipality being
         sampled.   The cleanup procedure  in Section 11 can  be used to
         overcome these interferences, but  unique samples may require
         additional  cleanup approaches to achieve the MDL listed in Table  1.
4.  Safety
    4.1  The toxicity or carcinogenicity  of each reagent used in this method
         has not been precisely defined;  however, each chemical  compound

-------
         should be treated as a potential  health  hazard.   From this view-
         point, exposure to these chemicals  must  be  reduced  to the lowest
         possible level  by whatever means  available.   The  laboratory  is
         responsible for maintaining a  current  awareness file of OSHA
         regulations regarding the safe handling  of  the chemicals specified
         in this method.  A reference file of material data  handling  sheets
         should also be made available  to  all personnel involved in the
         chemical  analysis.  Additional  references to  laboratory safety are
         available and have been identified3-5  for the information of the
         analyst.
5.  Apparatus and Materials
    5.1  Sampling equipment, for discrete  or composite sampling.
         5.1.1  Grab sample bottle - Amber Borosilicate or flint glass,
                1-liter  or 1-quart volume, fitted with screw caps lined with
                Teflon.   Aluminum foil  may be substituted  for Teflon  if the
                sample is not corrosive.   If amber bottles are not available,
                protect  samples from light.  The container  and cap liner must
                be washed, rinsed with  acetone  or methylene  chloride, and
                dried before use to minimize contamination.
         5.1.2  Automatic sampler (optional) -  Must  incorporate glass sample
                containers for the collection of  a minimum of 250 ml.
                Sample containers must  be  kept  refrigerated  at 4°C and
                protected from light during  compositing.   If the sampler
                uses a peristaltic pump, a minimum length  of compressible
                silicone rubber tubing  may be used.  Before  use, however,
                the compressible tubing should  be thoroughly rinsed with

-------
            methanol,  followed  by  repeated  rinsings  with distilled water
            to minimize the  potential  for contamination of the  sample.
            An integrating  flow meter  is  required  to collect  flow
            proportional  composites.
5.2  Glassware (All  specifications are suggested.  Catalog numbers are
     included for illustration  only.)
     5.2.1  Separatory funnel - 2000-mL,  with  Teflon stopcock.
     5.2.2  Drying column -  Chromatographic c  lumn 400 mm long  x 10 mm
                                            v'
            ID with coarse  frit.
     5.2.3  Chromatographic  column - 400 mm long x 19 mm ID with 250 ml
            reservoir  at the top and Teflon stopcock (Kontes  K-420290 or
            equivalent).
     5.2.4  Concentrator tube,  Kuderna-Danish  - 25-mL, graduated (Kontes
            K-570050-2525 or equivalent).  Calibration must be  checked
            at the volumes  employed in the test.   A  ground glass stopper
            is used  to prevent  evaporation of  extracts.
     5.2.5  Evaporative flask,  Kuderna-Danish  - 250-mL  (Kontes
            K-570001-0250 or equivalent).  Attach  to concentrator tube
            with springs.
     5.2.6  Snyder column,  Kuderna-Danish - three-ball macro  (Kontes
            K-503000-0121 or equivalent).
     5.2.7  Snyder column,  Kuderna-Danish - two-ball  micro (Kontes
            K-569001-0219 or equivalent).
     5.2.8  Vials -  Amber glass, 5 to  10 ml capacity with Teflon lined
            screw-cap.
     5.2.9  Volumetric flask -  2 ml with  glass stopper.

-------
    5.3  Boiling chips - approximately 10/40 mesh carborundum.   Heat  to
         400°C for 4 hours or extract  in  a Soxhlet extractor  with  methylene
         chloride.
                                                                  i
    5.4  Water bath - Heated, capable  of  temperature  control  (+2°C).   The
         bath should be used in a hood.
    5.5  Balance - Analytical, capable of accurately  weighing to the
         nearest 0.0001 g.
    5.6  Liquid chromatograph - Analytical  system complete  with  liquid
         chromatograph and all required accessories including syringes,
         analytical  columns, detector  and strip-chart recorder.  A data
         system is recommended for measuring peak areas.
         5.6.1  Pump - Isocratic pumping  system,  constant flow.
         5.6.2  Column 1 - Reversed-phase column, 5 micron  Spherisorb-ODS,
                250 x 4.6 mm or equivalent.
         5.6.3  Column 2 - Reversed-phase column, 5 micron  Lichrosorb  RP-2,
                250 x 4.6 mm or equivalent.
         5.6.4  Detector - Ultraviolet absorbance detector, 254  nm.
6.  Reagents
    6.1  Reagent water - Reagent water is defined as  a water  in which  an
         interferent is not observed at the  method detection limit of  each
         parameter of interest.
    6.2  Methyl ene chloride, acetone,  acetonitrile, distilled-in-glass
         quality or equivalent.
    6.3  Sodium sulfate (ACS) Granular, anhydrous; heated in a muffle
         furnace at 400°C overnight.
    6.4   Sodium  hydroxide,  IN -  Prepared  by  addina  4  a of sodium hvdrnxidp tn
        distilled water and  diluting to  100 ml.

-------
6.5  Sulfuric acid, IN - Prepared by diluting 2.8 ml of concentrated
     sulfuric acid to distilled water and diluting to 100 ml.
6.6  Florisil  - PR grade (60/100 mesh).   Purchase activated  at  1250°F
     and store in  brown  glass  bottle.  To prepare for use, place  150 g
     in a wide-mouth jar and heat overnight  at  160-170°C. Seal tightly
     with Teflon or aluminum foil-lined  screw cap and cool to room
     temperature.
6.7  Stock standard solutions  (1.00  ug/uL) - Stock standard  solutions
     can be prepared from pure standard  materials or purchased  as
     certified solutions.
     6.7.1  Prepare stock standard solutions by  accurately weighing  about
            0.010  grams  of pure material. Dissolve  the material  in
            distilled-in-glass quality acetonitrile  and dilute  to volume
            in a 10-mL volumetric  flask.   Larger  volumes  can be used at
            the convenience of the analyst.   If  compound  purity is
            certified  at 96% or greater,  the weight  can be used without
            correction to calculate  the  concentration of  the stock
            standard.  Commercially  prepared stock standards can  be  used
            at any concentration if  they  are certified by the manufacturer
            or by  an independent source.
     6.7.2  Transfer the stock standard  solutions into Teflon-sealed
            screw-cap  bottles.   Store at  4°C and  protect  from light.
            Stock  standard solutions should  be checked frequently for
            signs  of degradation or  evaporation,  especially just  prior
            to preparing calibration standards from  then.
     6.7.3  Stock  standard solutions must be replaced after six months
            or sooner  if comparison  with  check standards  indicates a
            problem.

-------
7.  Calibration
    7.1  Establish liquid chromatographic operating  parameters  equivalent  to
         those indicated in Table 1.   The liquid  chromatographic  system  can
         be calibrated using the external  standard technique  (Section  7.2)
         or the internal  standard technique  (Section 7.3).
    7.2  External  standard calibration  procedure:
         7.2.1  For each compound of  interest,  prepare  calibration  standards
                at a minimum of three concentration  levels  by adding volumes
                of one or more stock  standards  to a  volumetric  flask and
                diluting to volume with acetonitrile.   One  of the external
                standards should be at a concentration  near, but  above, the
                method detection limit.  The  other concentrations should
                correspond to the expected  range  of  concentrations  found in
                real  samples or should  define the working range of  the
                detector.
         7.2.2  Using injections of 2 to 5 uL of each calibration standard,
                tabulate peak height  or area  responses  against the  mass
                injected.  The results  can be used to prepare a calibration
                curve for each compound.  Alternatively, the ratio of the
                response  to the mass  injected,  defined  as the calibration
                factor (CF), can be calculated  for each compound  at each
                standard  concentration.  If the relative standard deviation
                of the calibration factor is  less than  10%  over the working
                range, linearity through the origin  can be  assumed  and the
                average calibration factor can  be used  in place of  a
                calibration curve.

-------
     7.2.3  The working calibration curve or calibration factor must  be
           •'verified on each working shift by the measurement  of one  or
            more calibration standards.   If the response for any
                                          /
            compound varies from the predicted response  by more than
            +10%, the test must be repeated  using a  fresh calibration
            standard.  Alternatively, a  new calibration  curve  or
            calibration factor must be prepared for  that compound.
7.3  Internal  standard calibration procedure.  To use this  approach,  the
     analyst must select one or more internal  standards  similar in
     analytical  behavior to the compounds of interest.   The analyst must
     further demonstrate that the measurement of the internal  standard
     is not  affected by method or matrix interferences.   Due to these
     limitations, no internal  standard applicable to all  samples can  be
     suggested.
     7.3.1  Prepare calibration standards at a minimum of three concen-
            tration levels for each compound of interest  by adding
            volumes of one or more stock standards to a  volumetric
            flask.   To each calibration  standard, add a  known  constant
            amount of one or more internal  standards, and dilute to
            volume with acetonitrile.  One of the standards  should be at
            a  concentration near, but above, the method  detection limit.
            The  other concentrations should  correspond to the  expected
            range of concentrations found in real  samples,  or  should
            define  the working range of  the  detector.
     7.3.2  Using injections of 2 to 5 uL of each calibration  standard,
            tabulate the peak  height or  area responses against  the

-------
            concentration for each compound and internal  standard.
            Calculate response factors (RF) for each compound as
            fol 1ows:
                          RF « (AsCis)/(AiSCs)
            where:
                 As  = Response for the compound to be measured.
                 A-J s = Response for the internal standard.
                 Cjs = Concentration of the internal  standard in ug/L.
                 Cs  = Concentration of the compound to be  measured in
                       ug/L.
            If the RF value over the working range is constant,  less
            than 10% relative standard deviation, the RF can  be  assumed
            to be invariant and the average RF can be used  for
            calculations.  Alternatively, the results can be  used to
            plot a calibration curve of response ratios,  As/AiS
            against RF.
     7.3.3  The working calibration curve or RF must be verified on each
            working shift by the measurement of one or more calibration
            standards.  If the response for any compound  varies  from the
            predicted response by more than 4-10%, the test  must  be
            repeated using a fresh calibration standard.  Alternatively,
            a  new calibration curve must  be prepared  for  that compound.
7.4  Before using any cleanup procedure,  the analyst  must process a
     series of calibration standards through the procedure  to validate
     elution patterns and the absence of  interferences from the
     reagents.

-------
8.  Quality Control
    8.1  Each laboratory using this method  is  required  to  operate a  formal
         quality control program.   The minimum requirements  of  this  program
         consist of  an  initial  demonstration of laboratory capability and
         the analysis of spiked samples as  a continuing check on performance.
         The laboratory is required to maintain performance  records  to define
         the quality of data that  is generated.
         8.1.1  Before  performing  any analyses, the  analyst  must demonstrate
                the  ability to generate acceptable accuracy  and precision
                with this method.   This ability is establishd as described
                in Section 8.2.
         8.1.2  In recognition of  the rapid advances occurring  in chromatog-
                graphy, the analyst is permitted certain options to  improve
                the  separations or lower the cost of measurements.   Each
                time such modifications to  the method are  made, the  analyst
                is required to repeat the procedure  in  Section  8.2.
         8.1.3  The  laboratory must spike and  analyze a minimum of 10% of
                all  samples to monitor continuing laboratory performance.
                This procedure is  described in Section  8.4.
    8.2  To establish the ability  to generate  acceptable accuracy and
         precision,  the analyst must perform the following operations.
         8.2.1  Select  a representative spike  concentration  for each
                compound to be measured. Using stock standards, prepare a
                quality control  check sample concentrate in methanol 1000
                times more concentrated than the selected  concentrations.

-------
     8.2.2  Using a pi pet, add 1.00 mL of the check sample concentrate
            to each of a minimum of four 1000-mL aliquots of reagent
            water.  A representative wastewater may be used in place of
            the reagent water, but one or more additional  aliquots must
            be analyzed to determine background levels,  and the spike
            level must exceed twice the background level  for the test to
            be valid.  Analyze the aliquots according  to the method
            beginning in Section 10.
     8.2.3  Calculate the average percent recovery (R),  and the standard
            deviation of the percent recovery (s), for the results.
            Wastewater background corrections must be  made before R and
            s calculations are performed.
     8.2.4  Using the appropriate data from Table 2, determine the
            recovery and single operator precision expected for the
            method, and compare these results to the values measured in
            Section 8.2.3.  If the data are not comparable, the analyst
            must review potential  problem areas and repeat the test.
8.3  The analyst must calculate method performance criteria and define
     the performance of the laboratory for each spike  concentration and
     parameter being measured.
     8.3.1  Calculate upper and lower control  limits for  method
            performance as follows:
                     Upper Control  Limit (UCL)  = R + 3 s
                     Lower Control  Limit (LCL)  = R - 3 s
            where R and s are calculated as in  Section 8.2.3.   The UCL
            and LCL can be used to  construct  control charts**  that  are
            useful  in observing trends in performance.

-------
     8.3.2  The laboratory must  develop  and  maintain  separate  accuracy
            statements  of laboratory  performance  for  wastewater  samples.
            An accuracy statement for the method  is defined  as R +_ s.
        i
            The accuracy statement  should be developed  by the  analysis
            of four aliquots of  wastewater as described in Section
            8.2.2,  followed  by the  calculation of R and s.   Alternately,
            the analyst may  use  four  wastewater data  points  gathered
            through the requirement for  continuing quality control  in
            Section 8.4. The accuracy statements should be  updated
            with this method. This ability  is established as  described
            regularly.
8.4  The laboratory is  required  to  collect in duplicate a portion of
     their samples  to monitor spike recoveries.  The  frequency of spiked
     sample analysis must be at  least 10% of all  samples or  one  sample
     per month, whichever is greater.  One aliquot of the sample must be
     spiked and analyzed as  described in Section  8.2.  If the  recovery
     for a particular compound does not  fall within the control  limits
     for method performance, the results reported for that compound in
     all samples processed  as part  of the same set must be qualified as
     described in Section 13.3.   The  laboratory should  monitor the
     frequency of data so qualified to ensure that it remains  at or
     below 5%.
8.5  Before processing any  samples, the  analyst should  demonstrate
     though the analysis of  a 1-liter aliquot of  reagent water that all
     glassware and  reagents  interferences are under control.  Each time

-------
         a set of samples is  extracted  or  there  is a  change  in  reagents, a
         laboratory reagent blank  should be  processed as a safeguard against
         laboratory contamination.
    8.6  It is recommended that  the  laboratory adopt  additional quality
         assurance practices  for use with  this method.  The  specific
         practices that  are most productive  depend upon the  needs of the
         laboratory and  the nature of the  samples.  Field duplicates may be
         analyzed to monitor  the precision of the sampling technique.  When
         doubt exists over the identification of a peak on the  chromatogram,
         confirmatory techniques such as liquid chromatography  with a
         dissimilar column, must be  used.  Whenever possible, the laboratory
         should perform  analysis of  standard reference materials and
         participate in  relevant performance evaluation studies.
9.  Samples Collection,  Preservation, and  Handling
    9.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices7  should  be  followed; however, the bottle must
         not be prerinsed with sample before collection.  Composite samples
         should be collected  in  glass containers in accordance  with the
         requirements of the  program.  Automatic sampling equipment must
         be as free as possible  of Tygon and other potential sources of
         contamination.
    9.2  The samples must be  iced  or refrigerated at  4°C from the time of
         collection until extraction.
    9.3  Adjust the pH of the sample to 6  to 8 with IN sodium hydroxide
         or IN sulfuric acid  immediately after sampling.

-------
    9.4  All  samples must be extracted  within  seven  days  and  completely
         analyzed  within  40  days  of extraction.
10.  Sample Extraction
    10.1 Mark the  water meniscus  on the side of the  sample bottle  for  later
         determination of sample  volume.   Pour the entire sample into  a
         2-liter separatory  funnel.  Check the pH of the  sample with wide
         range pH  paper and  adjust  to 7 with 1 N sodium hydroxide  or 1 N
         H2S04.
    10.2 Add  60 mL of methylene chloride to the sample bottle, seal, and
         shake 30  seconds to rinse  the  inner walls.  Transfer the  solvent to
         the  separatory funnel and  extract the sample by  shaking the funnel
         for  2 min with periodic  venting  to release excess pressure.  Allow
         the  organic layer to separate  from the water phase for a minimum of
         10 min.   If the  emulsion interface between layers is more than one
         third the volume of the  solvent  layer, the analyst must employ
         mechanical  techniques to complete the phase separation.  The
         optimum technique depends  upon the sample, but may include
         stirring,  filtration of the  emulsion through glass wool,
         centrifugation,  or  other physical methods.  Collect the methylene
         chloride  extract  in  a 250-mL Erlenmeyer flask.
    10.3 Add  a second  60-mL  volume  of methylene chloride  to the sample
         bottle and  repeat the extraction procedure a second time,
         collecting  the extract.  Perform a third extraction in the same
         manner and  combine  the extracts.

-------
10.4 Assemble a Kuderna-Danish (K-D)  concentrator  by  attaching  a  25-mL
     concentrator tube to a  250-mL  evaporative  flask.   Other concentra-
     tion devices or techniques may be used  in  place  of the K-D if  the
     requirements of Section 8.2 are  met.
10.5 Pour the combined extract through a drying column  containing about
     10 cm of anhydrous sodium sulfate, and  collect the extract in  the
     K-D concentrator.  Rinse the Erlenmeyer flask and  column with  20 to
     30 ml  of methylene chloride to complete the quantitative transfer.
     Once the flask rinse has passed  through the drying column, rinse
     the column with 30 to 40 ml of methylene chloride.
10.6 Add 1 or 2 clean boiling chips to the evaporative  flask and  attach
     a three-ball  Snyder column.  Prewet the Snyder column by adding
     about 1 mL methylene chloride  to the  top.   Place the K-D apparatus
     on a hot water bath, 60 to 65°C, so that the  concentrator  tube is
     partially immersed in the hot  water,  and the  entire lower  rounded
     surface of the flask is bathed with hot vapor.   Adjust the vertical
     position of the apparatus and  the water temperature as required to
     complete the concentration in  15 to 20  min.   At  the proper rate of
     distillation, the balls of the column will  actively chatter  but the
     chambers will  not flood with condensed  solvent.  When the  apparent
     volume of liquid reaches 1 ml, remove the  K-D apparatus and  allow
     it to drain and cool for at least 10  minutes.
10.7 Remove the macro-Snyder column and rinse the  flask and its lower
     joint into the concentrator tube with 1 to 2  ml  of methylene
     chloride.  A 5-mL syringe is recommended for  this  operation.   Add 1

-------
          or 2 clean boiling chips and attach a two-ball  micro-Snyder column
          to the concentrator tube.  Prewet the micro-Snyder column  with
          methylene chloride and concentrate the solvent  extract  as  before.
          When an apparent volume of 0.5 ml is reached, or  the  solution stops
          boiling, remove the K-D apparatus and allow it  to drain and cool
          for 10 minutes.
     10.8 Remove the micro-Snyder column and adjust  the volume  of the extract
          to 1.0 ml with methylene chloride.  Stopper the concentrator tube
          and store refrigerated if further processing will  not be performed
          immediately.  If the extract is to be stored longer than two days,
          transfer the extract to a screw capped vial  with  a Teflon-lined
          cap.  If the sample extract requires no further cleanup, proceed
          with solvent exchange to acetonitrile as described beginning
          with Section 11.5.  If the sample requires  cleanup, proceed to
          Section 11.
    10.9 Determine the original  sample volume  by refilling the  sample  bottle
         to  the mark and transferring the water  to a  1000-mL graduated
         cylinder.  Record the sample volume to  the nearest  5 ml.
11.  Cleanup  and Separation
    11.1 Cleanup procedures may  not be necessary for  a relatively clean
         sample matrix.   The cleanup procedure recommended in this method
         has been used for the analysis of various clean  waters and
         industrial  effluents.   If particular  circumstances demand the use
         of  additional  cleanup,  the analyst must demonstrate that the
         recovery of each compound of interest  is  no  less than  85%.

-------
11.2 Slurry 10 g of Florisil in 100 ml of methyl ene chloride which has
     been saturated with reagent water.  Transfer the slurry to a
     chromatographic column (Florisil  is retained with a plug of glass
     wool).  Wash the column with 100 ml of methylene chloride.  Use a
     column flow rate of 2 to 2.5 mL/min throughout the wash and elution
     profiles.
11.3 Add the extract from Section 10.8 to the head of the column.  Allow
     the solvent to elute from the column until  the Flor*>i1  is almost
                                                        ., /
     exposed to the air.  Elute the column with 50 ml of methylene
     chloride.  Discard this fraction.
11.4 Elute the column with 50 ml of 5% acetone in methylene chloride.
     Collect this fraction in a K-D apparatus.  Concentrate the column
     fraction to 1 ml as described in Sections 10.6 and 10.7.
11.5 Add 15 ml of acetonitrile to the concentrate along with 1  or 2
     clean boiling chips.  Attach a three-ball micro-Snyder column to
     the concentrator tube.  Prewet the micro-Snyder column with
     acetonitrile and concentrate the solvent extract to an apparent
     volume of 1 mL.  Allow the K-D apparatus to  drain and  cool  for 10
     minutes.
11.6 Transfer the liquid to a 2-mL volumetric flask and dilute  to the
     mark with acetonitrile.  Mix thoroughly prior to analysis.   If the
     extracts will  not be analyzed immediately, they should be
     transferred to Teflon sealed screw-cap vials and refrigerated.
     Proceed with the liquid chromatographic analysis.

-------
12.  Liquid Chromatography
    12.1 Table 1 summarizes  the  recommended  operating  conditions  for the
         liquid chrcmatograph.   Included  in  this  table are  estimated
         retention  times  and method  detection  limits that can be  achieved by
         this method.   An example of the  separations achieved by  Column 1
         and Column 2  are shown  in Figures  1  and  2.  Other  columns,
         chromatographic  conditions, or detectors may  be  used if  the
         requirements  of  Section 8.2 are  met.
    12.2 Calibrate  the liquid chrcmatographic  system daily  as described in
         Section 7.
    12.3 If an internal standard approach is being  used,  the analyst
         must not add  the internal standard to sample  extracts  until
         immediately before injection into  the instrument.
    12.4 Inject 2 to 5 uL of the sample extract by completely filling the
         sample valve  loop.   Record  the resulting peak sizes  in area or peak
         height units.
    12.5 The width  of  the retention  time  window used to make  identifications
         should be  based  upon measurements  of  actual retention  time
         variations of standards over the course  of a  day.   Three times the
         standard deviation  of a retention  time for a  compound  can be used
         to calculate  a suggested window  size; however, the  experience of
         the analyst should  weigh heavily in the  interpretation of
         chromatograms.

-------
    12.6 If the response for the peak  exceeds  the working  range of the
         system, dilute  the extract  and  reanalyze.
    12.7 If the measurement of the peak  response is prevented  by the
         presence of interferences,  further cleanup is  required.
13.  Calculations
    13.1 Determine the concentration of  individual  compounds  in the sample.
         13.1.1 If the external  standard calibration procedure is used,
                calculate the amount of  material  injected  from the peak
                response using the calibration curve or calibration factor
                in Section 7.2.2.   The concentration in the  sample can be
                calculated as follows:
                                            (A)(Vt)
                     Concentration,  ug/L =

                where:
                   A  -  Amount of material  injected, in nanograms.
                   Vi  =  Volume of extract injected in ul_.
                   Vt »  Volume of total  extract in uL.
                   Vs =  Volume of water  extracted in mL.
         13.1.2 If the internal  standard calibration procedure was used,
                calculate the concentration in the sample  using the response
                factor (RF) determined in Section 7.3.2 as follows:
                                              (As)(Is)
                     Concentration,  ug/L =  (A.$)(RF)(VQ)
                'where:
                   As = Response for the compound to be measured.

-------
                   A-JS = Response for the internal  standard.
                   Is  » Anount of internal  standard  added  to  each
                         extract in ug.
                   V0  " Volume of water extracted, in  liters.
    13.2 Report results in micrograms per liter without correction  for
         recovery data.  When duplicate and  spiked  samples  are  analyzed,
         report all  data obtained with the sample results.
    13.3 For samples  processed as part of a  set  where the laboratory  spiked
         sample recovery falls outside of the control limits in Section 8.3,
         data for the affected compounds must be labeled as suspect.
14.  Method Performance
                                         Q
    14.1 The method detection limit (MDL)   is defined as the minimum
         concentration of a substance that can be measured  and  reported with
         99% confidence that  the value is  above  zero.   The  MDL  concentra-
         tions listed in Table 1 were obtained using  reagent water.1
         Similar results were achieved using  representative wastewaters.
    14.2 This method  has been tested  for linearity  of recovery  from spiked
         reagent water and has been demonstrated  to be  applicable over the
         concentration range  from 10  x MDL to 1000  x MDL.
    14.3 In  a single  laboratory,  Battelle  Columbus  Laboratories, using
         spiked wastewater samples, the average  recoveries  presented  in
         Table 2 were obtained.   Seven  replicates of each of two different
         wastewaters  were spiked and  analyzed.   The standard deviation of
         the percent  recovery is  also  included in Table 2.1

-------
                                 REFERENCES
1.  "Development of Methods for Pesticides in Wastewaters," Report for EPA
    Contract 68-03-2956 (In preparation).

2.  ASTM Annual Book of Standards, Part 31, D3694, "Standard Practice for
    Preparation of Sample Containers and for Preservation," American Society
    for Testing and Materials, Philadelphia, PA, p. 679, 1980.

3.  "Carcinogens - Working with Carcinogens," Department of Health, Education,
    and Welfare, Public Health Service, Center for Disease Control, National
    Institute for Occupational Safety and Health, Publication No. 77-206,
    August, 1977.

4.  "OSHA Safety and Health Standards, General  Industry," (29 CFR 191.0),
    Occupational Safety and Health Administration, OSHA 2206 (Revised,
    January, 1976).

5.  "Safety in Academic Chemistry Laboratories," American Chemical  Society
    Publication, Committee on Chemical Safety,  3rd Edition, 1979.

6.  "Handbook for Analytical  Quality Control  in Water and Wastewater
    Laboratories," EPA-600/4-79-019, U. S. Environmental Protection Agency,
    Environmental Monitoring and Support Laboratory - Cincinnati, Ohio 45268,
    March, 1979.

7.  ASTM Annual Book of Standards, Part 31, D3370, "Standard Practice for
    Sampling Water," American Society for Testing and Materials, Philadelphia,
    PA, p. 76, 1980.

8.  Glaser, J. A. et al, "Trace Analysis for Wastewaters," Environmental
    Science and Technology, 15, 1426 (1981).

-------
       TABLE  1.   CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
                       Retention Time (Mln)                  MDL
  Parameter            Column 1Column  2                  (u§/L)


Oryzalin                 6.2       10.7                     0.5


Column 1 conditions:  Spherisorb-ODS, 5 micron,  250  x  4.6 mm;  1 mL/min  flow;
40/60 acetonitrile/water.  A UV detector  was  used  with this  column to determine
the MDL.

Column 2 conditions:  Lichrosorb RP-2,  5  micron, 250 x 4.6 mm;  1 mL/min  flow;
50/50 acetronitrile/water.

-------
               TABLE 2.   SINGLE  LABORATORY ACCURACY AND  PRECISION^)

Parameter
Dry za Tin


Sample
Type(b)
1
2

Background,
yg/L
4
40
Spike
Level
ug/L
10
200
Mean
Recovery
(X)
106
100
Standard
Deviation
(X)
6
10
Number
of
Replicates
7
7
(a)   Column  1  conditions  were  used.
(b)   1 = Relevant  industrial wastewater diluted 1000:1 with municipal sewage
         effluent;
     2 » Relevant  industrial wastewater diluted 100:1 with municipal sewage
         effluent.

-------
                         Oryzalin
                                                                                        1
 I  » ' ' '  |  '  ' ' ' I  '  ' ' ' I  '  « ' * I  '  ' ' ' I  '  ' ' ' I  '  ' '
6.0      8.0     10.0      12.0     H.O     16.0     18.0   20.0

    Retention Time,  Min.
I  I  1 I | I  I  I ' I'
  .   2.0      4.0
              FIGURE 1.  HPLC-UV CHROMATOGRAM OF 10 ng OF ORYZALIN  (COLUMN 1).

-------
                                    Oryzalin
                      8.0      10.0     12.0      14.0      16.0    18.0   20.0
               Retention Time, Min.





FIGURE 2.   HPLC-UV CHROMATOGRAM  OF 250 ng OF ORYZALIN (COLUMN 2).

-------