EPA-600/4-78-012
                                                 January 1978
METHODS FOR MEASURING THE ACUTE  TOXICITY OF EFFLUENTS
                TO AQUATIC  ORGANISMS
                          by
                   William  Peltier
                Bioassay  Subcommittee
          EPA Biological  Advisory Committee
          Environmental Research Laboratory
                Athens, Georgia  30601
                              U.S. Environmental Protection Agency
                              Region 5, Library (PL-12J)
                              77 West Jackson Boulevard. 12th Floor
                              Chicago, IL  60604-3590
                   Project  Officer

                 Cornelius  I.  Weber
               Aquatic  Biology Section
    Environmental Monitoring  & Support Laboratory
               Cincinnati,  Ohio 45268
    ENVIRONMENTAL MONITORING & SUPPORT LABORATORY
         OFFICE OF RESEARCH  AND DEVELOPMENT
         U.S. ENVIRONMENTAL  PROTECTION AGENCY
               CINCINNATI, OHIO 45268

-------
                                DISCLAIMER
                                     V

     This report has been reviewed by the Environmental Monitoring and
Support Laboratory, U.S. Environmental Protection Agency,  and approved
for publication.  The mention of trade names or commercial products does
not constitute endorsement or recommendation for use.
                                      11

-------
                                FOREWORD

    Environmental measurements are required to determine the quality of
ambient water, the character of effluents, and the effects of pollutants
on aquatic life.  The Environmental Monitoring and Support Laboratory-
Cincinnati conducts research to develop, evaluate, and promulgate methods
to:

       Measure the presence and concentration of physical, chemical and
       radiological pollutants in water, wastewater, bottom sediments,
       and solid waste.
       Concentrate, recover, and identify enteric viruses, bacteria, and
       other microorganisms in water.
       Measure the effects of pollution on freshwater, estuarine, and
       marine organisms, including the phytoplankton, zooplankton, peri-
       phyton, macrophyton, macroinvertebrates, and fish.
       Automate the measurement of the physical, chemical, and biological
       quality of water.
       Conduct an Agency-wide quality assurance program to assure stand-
       ardization and quality control of systems for monitoring water and
       wastewater.

    The Federal Water Pollution Control Act Amendments of 1972 (PL 92-500)
explicitly state that it is the national policy that the discharge of
toxic substances in toxic amounts be prohibited.  Determination of the
toxicity of effluents, therefore, has high priority in the EPA water pol-
lution control program.  However, suitable, standardized methodology
for effluent bioassays has not been available to EPA regional and state
programs.  This report fills an urgent current need for standardized
methods to measure the toxicity of effluents to aquatic life.
                                    Dwight G. Ballinger
                                    Director
                                    Environmental Monitoring and
                                    Support Laboratory - Cincinnati
                                     iii

-------
                                 PREFACE
    The effluent toxicity tests described in this document were prepared
by the Bioassay Subcommittee of the EPA Biological Advisory Committee
to provide methods needed for effluent monitoring by EPA Regional and
State NPDES programs and for self-monitoring by NPDES permit holders.
    Two types of toxicity tests are described:

    1.  A preliminary, short-term, static, range-finding test for use in
        determining the concentrations of effluent to be used in a "defini-
        tive" test.

    2.  A long-term (generally 96-hr), flow-through, definitive test for
        use in determining the acute toxicity of the effluent, expressed
        as a LC50 or EC50.

     These methods will be  included in the second edition of the EPA manual,
"Biological Field and Laboratory Methods for Measuring the Quality of
Surface Waters and Effluents," but were printed separately in limited
quantity to make them available prior to the publication of the manual.

                      Bioassay Subcommittee Members

    William Peltier, Chairman, S&A Division, Region IV
    John Arthur, Environmental Research Laboratory, Duluth
    Bruce Binkley, National Enforcement Investigation Center, Denver
    Stanley Hegre, Environmental Research Laboratory, Narragansett
    William Horning, Newtown Fish Toxicology Laboratory
    Philip Lewis, Environmental Monitoring & Support Lab, Cincinnati
    Royal Nadeau, S&A Division, Region II
    Ronald Preston, S&A Division, Region III
    Steven Schimmel, Environmental Research Laboratory, Gulf  Breeze
    Milton Tunzi, S&A Division, Region IX

                              Contributors

    Ronald Eisler, Environmental Research Lab, Narragansett
    Charles Stephan,    "          "       "  , Duluth
    Lee Tebo, S&A Division, Region IV
    Cornelius I. Weber, Environmental Monitoring &  Support Lab., Cincinnati
                                Cornelius I. Weber, Ph.D.
                                Chairman, Biological Advisory Committee
                                Chief, Aquatic Biology Section
                                Biological Methods Branch
                                Environmental Monitoring & Support Laboratory
                                      IV

-------
                                ABSTRACT

     This report describes methods for the measurement of the acute
toxicity of effluents to macroinvertebrates and fish.  The methods include
a preliminary short-term (8-24 hr.)» range-finding test and a
long term (96 hr.), flow-through (or alternate static) definitive test for
use in determining the LC50 or EC50 of the effluent.  The report includes
guidelines on effluent sampling and holding, facilities and equipment,
dilution water, test species selection and handling, and data interpretation.

-------
                               CONTENTS

Foreword	
Preface	     iv
Abstract 	      v
Figures	   viii
Tables	     ix

   1.  Introduction  	      1
   2.  Facilities and Equipment  	      4
           General Requirements  	      4
           Construction Materials  	      4
           Effluent (Toxicant) Delivery System 	      5
           Test Chambers	      6
               Type	      6
               Cleaning  	      6
   3.   Test Organisms	      7
           Species 	      7
           Source	      7
           Size	      7
           Holding and Handling Care	     10
           Disease Treatment  	     10
           Transportation and Acclimation  	     13
   4.  Dilution Water	     14
   5.  Effluent Sampling and Holding 	     17
           Sampling	     17
           Holding   	     18
   6.  Test Procedure	     19
           Range-Finding Test	     19
           Definitive Test	     19
               Test Conditions	     19
               Number of Test Organisms	     20
               Loading of Test Organisms	     20
               Test temperature	     20
               Oxygen requirements and aeration	     21
               Beginning the  test	     21
               Feeding	     21
               Duration	     22
   7.  Test Results	     22
               Biological data	     22
               Chemical and physical data	     22
               Calculation of LC50 and EC50	     24
               Reports	     24

References	     26
Appendices

   A.  Litchfield and Wilcoxon abbreviated method  	     29
   B.  Log-concentration versus percent survival method  ....     37
   C.  Dilutor systems, control panel and equipment lists  ...     39
                                     vii

-------
                                 FIGURES

Number                                                           page

  1     Data Sheet for Effluent Toxicity Test	    33

  2     Line fitted to data, and LC16, LC50 and LC84 as read
        from the line	   34
                                   2
  3     Nomograph for obtaining Chi  from expected-percent-
        affected and observed-percent-affected minus the
        expected-percent-affected .... 	   35

  4     Nomograph for raising base S to a fractional exponent .   36

  5     Plotted data and fitted line for log-concentration
        versus percent survival method  	   38

  6     Photographs of dilutor systems. A.  Solenoid valve
        dilutor system.  B.  Vacuum siphon dilutor system ....   40

  7     Solenoid valve dilutor system; general diagram  ....   41

  8     Solenoid valve dilutor system; detailed diagram ....   42

  9     Vacuum siphon dilutor system;  general diagram 	   44

 10     Vacuum siphon dilutor system;  detailed diagram  ....   45

 11     Dilution water chambers 	   47

 12     Effluent chambers 	   48

 13     Mixing chamber	   49

 14     Dilutor control panel 	   50
                                    viii

-------
                                 TABLES

Number                                                             Page

  1     Recommended species and test temperatures 	     8

  2     Recommended prophylatic and therapeutic treatments
        for freshwater fish	    11

  3     Preparation of reconstituted fresh waters 	    15

  4     Preparation of reconstituted sea water  	    16

  5     Percentage of un-ionized ammonia in distilled water
        at various temperatures and pH's	    23

  6     Salinity correction factor  	    23

  7     Corrected values for 0% or 100% effect	    32

  8     Values of Chi   	    32
                                      IX

-------
                                SECTION 1

                              INTRODUCTION

     The Declaration of Goals and Policy of the Federal Water Pollution
Control Act Amendments of 1972,  Section 101(a)(3), states that "it is
the national goal that the discharge of toxic pollutants in toxic amounts
be prohibited".  Current Agency programs for the protection of aquatic
life in receiving waters are based in part on effluent limitations for
individual chemicals.  However,  toxicity data are available for only a
limited number of compounds.  The effluent limitations, therefore, may not
provide adequate protection where the toxicity of the components in the
effluent is not known, where there are synergistic effects between toxic
substances in complex effluents, and/or where a complete chemical charac-
terization of the effluent has not been carried out.  Since it is not
economically feasible to determine the toxicity of each of the thousands
of potentially toxic substances in complex effluents or to conduct
an exhaustive chemical analysis of the effluent, the most direct and cost-
effective approach to the measurement of the toxicity of effluents is to
conduct a bioassay with aquatic organisms representative of indigeneous
populations.  For this reason, the use of effluent bioassays to identify
and control toxic discharges is rapidly increasing within the Agency and
state NPDES programs.

     The lack of standardized bioassay methodology developed explicitly
for effluents has delayed the implementation of EPA regional and state
effluent toxicity testing programs and has resulted in a lack of uniformity
in test procedures.  In response to this problem, a subcommittee was
organized within the EPA Aquatic Biology Methods Advisory Committee,
sponsored by the Environmental Monitoring and Support Laboratory, Cincinnati,
to prepare effluent bioassay methods for the NPDES program.  To provide
valid methods, it was essential to include EPA Regional and Enforcement
programs personnel with extensive experience in conducting effluent toxi-
city tests.  In completing their task, the subcommittee members drew from
their own experience and borrowed heavily from the report, "Methods for
Acute Toxicity Tests with Fish,  Macroinvertebrates and Amphibians,"
(EPA 660/3-75-009), previously prepared primarily to standardize basic
laboratory methods for determining the toxicity of pure compounds.

     The acute toxicity tests for effluents described in this report are
used to determine the effluent concentration, expressed as a percent volume,
that is lethal to 50 percent of the organisms within 96 hrs or some other
prescribed period of time.  This value is termed the median lethal concen-
tration or LC50.  Where death is not easily detected, for example, with
some invertebrates, other indicators such as immobilization must be used
as the "adverse effect." Blood chemistry, biochemical measurements or
histological examinations can also be employed to measure the toxicity of
an effluent.  The concentration of effluent, expressed as a percent
volume, that causes such a defined adverse effect (other than death) in
50 percent of the test organisms within the prescribed exposure period is
termed the median effective concentration or EC50.

-------
     A two-test sequence is generally required to estimate the acute
toxicity of an effluent:  (1) a preliminary, short-term (8-24 hr),
range-finding test is conducted to define the range of effluent dilutions
to be used in the definitive test, and (2) a more rigorous long-term
definitive test is conducted (using the range of effluent dilutions
determined by the range-finding test) over a 96-hour time period to arrive
at the acute toxicity of the effluent which is expressed as a LC50 or EC50.

     One of the following procedures shall be used for the range-finding
test:

     A.  Test organisms are placed in suitable containers and exposed
         under static conditions to 3-5 widely-spaced dilutions of the
         effluent for a period of 8-24 hours.

     B.  If the effluent has a high dissolved oxygen demand, or the
         organisms require flowing water, a flow-through test is used
         to define the range of toxicity of the effluent.

     One of the following procedures shall be used for the definitive test:

     A.  Preferred procedures (flow-through tests)

         1.  Test organisms are exposed to effluent solutions flowing
             into and out of test chambers on a once-through basis for
             the duration of the test.  The effluent is conveyed directly
             and continuously from the source to the dilutor system.

         2.  Test organisms are exposed to effluent solutions flowing into
             and out of test chambers on a once-through basis for the
             duration of the test as in A.I above.  However, the effluent
             is supplied to the dilutor system from discrete effluent samples
             collected periodically.  The interval at which samples are
             collected is based on the variability of the effluent charac-
             teristics, production schedule, batch processes, retention
             time, etc.  (see p. 17  - Effluent Sampling and Holding).

     The continuous effluent sampling technique (A.I) is the best of two
methods described.

     B.  Alternative procedures (static tests)

         1.  Test organisms are exposed to a fresh solution of the same
             concentration of effluent every 24 hours, either by trans-
             ferring the test organisms from one test chamber to another,
             or by replacing the effluent solution in the test chambers.

         2.  Test organisms are exposed to the same effluent solution for
             the duration of the test.  However, the effluent solution in

-------
         each test chamber may be filtered, aerated,  or sterlized by
         continuous circulation through an appropriate apparatus.

     3.  Test organisms are exposed to the original effluent solution
         for the duration of the test without being continuously
         circulated through an apparatus as in B.2.

     The alternative procedures may be used in emergency situations or
where adequate facilities are not available to the investigator.  However,
it must be understood that it is not a valid test unless it can be con-
clusively demonstrated that the chemical characteristics and toxicity of
the effluent do not change over time.  Because of toxicant adsorption on
the test chambers, uptake by test organisms and the effect of metabolites
on toxicity, it is preferred that the effluent solution be renewed at
least once every 24 hours as described in B.I above.
     The special environmental requirements of some organisms, such as
flowing water, fluctuating water levels, or substrate may preclude the use
of static testing.

-------
                                SECTION 2

                        FACILITIES AND EQUIPMENT
GENERAL REQUIREMENTS

     Effluent toxicity tests may be carried out in a fixed or mobile lab.
Depending upon the scope of the bioassay program, facilities may include
equipment for rearing, holding and acclimating organisms.  Temperature
control is achieved using circulating water baths, or environmental
chambers.  Dilution water may be ground water, surface water, reconstituted
water, or dechlorinated tap water.  Holding, acclimation, and dilution
water should be temperature controlled and aerated whenever possible.
Air used for aeration must be free of oil and fumes; filters to remove
oil in water are desirable.  Test facilities must be well ventilated and
free of fumes.  During holding, acclimating, and testing, test: organisms
should be shielded from disturbances.

     Some organisms may have special environmental requirements such as
f lowing-^water, fluctuating water levels, or substrate which must be
provided.  During holding, acclimating, and testing, immature stream
insects should always be in flowing water as described by Nebeker and
Lemke (1968); penaeid shrimp and bottom-dwelling fish should be provided
a silica sand substrate.  Since cannibalism can occur among many species
of arthropods, they should be isolated by some means (e.g., with screened
compartments), or the claws of crabs and crayfish should be bound.

CONSTRUCTION MATERIALS

     Materials used in the construction of the test equipment which
come in contact with the effluent should be carefully chosen.  Glass,
No. 316 stainless steel, and perfluorocarbon plastics (TEFLONR) should
be used whenever possible to minimize leaching, dissolution, and sorption.
Linear polyethylene may also be used with some types of  industrial
and municipal effluents, but should be avoided with those containing
synthetic organic compounds or pesticides.  Unplasticized plastics such as
polyethylene, polypropylene, TYGONR and fiberglass can be used for holding,
acclimating,  and dilution-water storage tanks, and in the water delivery
system. Copper, galvanized material, rubber, brass, and  lead must not come
in contact with holding, acclimation, or dilution water, or with effluent
samples and  test solutions.

-------
EFFLUENT (TOXICANT) DELIVERY SYSTEM (Flow-through test only)

     Although many types of toxicant delivery systems have been designed*,
the flow-through, proportional-dilutor delivery system has proven to be
the best and the preferred system for routine effluent toxicity tests
conducted in, both fixed and mobile laboratories.

     The following factors should be considered in designing the system:
(1) whether the apparatus will be installed and used in a fixed or mobile
laboratory; (2) the existence of adequate space and/or structural require-
ments for the delivery system, test chambers, effluent and dilution-water
storage; (3) the applicability of the delivery system to specific effluent
characteristics (high suspended solids, volatiles, etc.); (4) the system's
dependability, durability, flexibility, and ease of maintenance and replace-
ment; (5) the ability to perform within the flow rate and concentration
accuracy limitations; and (6) the cost of the system.

     Two types of proportional dilutors are described in the appendix.  The
solenoid value system is preferred, but the vacuum siphon system is
acceptable, if funds are limited.

     The flow rate through the proportional dilutor must provide for
at least five complete water volume changes in 24 hours in each test
chamber, plus sufficient flow to maintain an adequate concentration of
dissolved oxygen.  It is often desirable to construct the dilutor with
an additional reserve flow capacity, depending on its application and/or
special effluent characteristics.  The flow rates through the test
chamber should not vary by more than 10 percent among test chambers at
any time during any test.  The dilutor should also be capable of maintain-
ing the test concentration in each test chamber within 5 percent of the
starting concentration for the duration of the test.

     The calibration of the dilutor should be checked carefully before
and after each test.  This check should determine the volume of effluent
and dilution water used in each portion of the effluent delivery system
and the flow rate through each test chamber.  The general operation of
the dilutor should be checked at least at the beginning and end of each day
during the test.
*(Lowe, 1964; Mount and Brungs, 1967; Sprague, 1969; Freeman, 1971; Cline
 and Post, 1972; Granmo and Kollberg, 1972; Bengtsson, 1972; Lichatowich,
ej^ al^. , 1973; Schumway and Palensky, 1973; Abram, 1973; Schimmel, Hansen,
and Forester, 1974; DeFoe, 1975; Riley, 1975).

-------
TEST CHAMBERS

Type

     Test chambers used in flow-through tests are constructed of 1/4 inch
plate glass held together with (GE) clear silicone adhesive.  Silicone
adhesive absorbs some organochlorine and organophosphorus pesticides which
are then difficult to remove.  Therefore, as little of the adhesive as
possible should be in contact with water; extra beads of adhesive inside
the containers should be removed.  Stainless steel (#316) can be used
in the construction of test chambers, but must be of welded, not soldered,
construction.

     The size of the chambers may vary according to the size of the test
organism and/or the facilities, but the test solution should have a mini-
mum depth of 5 cm.  All chambers should have either a glass or screen
cover to prevent organisms from jumping out.

     The test chambers most commonly used in static tests are wide-mouth,
3.9 liter (1-gallon) or 19.6 liter (5-gallon) soft-glass bottles.  Con-
tainers such as 1-liter battery jars or 250-ml beakers are often more
suitable as test chambers for fish eggs and/or larvae and small Crustacea.
Special glass or stainless steel test chambers can be constructed to
accommodate a test organism which requires special physical conditions.

Cleaning

     All test chambers, whether new or used, must be washed in the follow-
ing manner to remove surface contaminants:

     A.  Soak and wash in suitable detergent in water preferably heated to
         a temperature of 50°C or higher.  The detergent (powder or liquid)
         should not be a fatty-acid base but entirely synthetic (SPARKLEENR
         or ALCONOXR).

     B.  Rinse with water (preferably heated to 50°C or higher).

     C.  Rinse with a fresh, dilute (5 percent) hydrochloric acid for
         removing metals and bases.

     D.  Rinse with water (preferably heated to 50°C or higher).

     E.  Rinse with acetone to remove organic compounds.  When contaminated
         with a pesticide, test chambers must be rinsed with acetone
         before they are placed in the hot detergent soak (Item A. above).

     F.  Rinse twice with water.

     When feasible, the outlined cleaning procedure must be used for other
equipment that comes in contact with the dilutor system, pumps, tanks, etc.
All test chambers and equipment must be thoroughly rinsed with the dilution
water prior to each test.

-------
                                 SECTION  3

                              TEST ORGANISMS
 SPECIES

     Whenever possible, effluent  toxicity  tests should be  conducted with a
 sensitive  species  that  is  indigenous  to  the receiving water, readily
 available,  and either commercially or recreationally important.  Acceptable
 species  include those listed  in Table 1, which are sensitive to most
 toxicants.  The test organisms must be identified to species.

 SOURCE

     Although effluent  tests  should be conducted with species that are
 indigenous  to, or  stocked  into, the receiving water, the test organisms
 do not have to be  taken from  the  receiving water.  It is often difficult
 to obtain  fish of  the desired size and condition from the  receiving
 water.   Collection permits are often  difficult to obtain,  and the
 organisms  in the receiving water  may  have been previously  exposed to the
 effluent.   Fish captured by electroshocking must not be used in testing.
 The usual  sources  of freshwater fish  used for toxicity tests are private,
 state, and  Federal hatcheries.  If trout  are used as test organisms,
 they should be obtained from  stock that has been certified as disease-
 free.

     Some  fish, such as Fathead and Sheepshead minnows, are easily reared
 under laboratory conditions (USEPA, 1972; Schimmel and Hansen, 1974).
 However, it is more practical to  collect most marine species from the
 indigenous  population.

     Some  invertebrates such as daphnids, midges, and shrimp may be
 reared in  laboratory cultures.  Care must be taken to insure that only
 young age  groups and early instars are used in testing.  Daphnids from
 cultures in which  ephippia are being produced should not be used in the
 tests.   Invertebrates not amenable to laboratory rearing are usually
 obtained directly  from wild populations.

 SIZE

     Very immature fish (not actively feeding on exogenous food), spawning
 fish, or recently  spent fish should not be used.   Fish weighing between
 0.5 and  5.0 grams  each  are preferred.   In any single test, all fish
 should be taken from the same year class, and the total length (tip of
 snout to end of tail) of the longest fish should be no more than 1-1/2
 times that of the  shortest one.    Immature invertebrates should be used
whenever possible.

-------
            TABLE 1.   RECOMMENDED SPECIES AND TEST TEMPERATURES
                                                                      o
Species                                          Test temperature (°C)
Freshwater
   Vertebrates
      Coho salmon, Oncorhynchus kisutch                       12
      Rainbow trout, Salmo gairdneri                          12
      Brook trout, Salvelinus fontinalis                      12

      Goldfish, Carassius auratus                             22
      Fathead minnow, Pimephales promelas                     22
      Channel catfish, Ictalurus punctatus                    22
      Bluegill, Lepomis macrochirus                           22

                o
   Invertebrates
      Daphnids, Daphnia magna or I), pulex                     17
      Amphipods, Gammarus lacustris, (2. fasciatus, or         17
                 G_. pseudolimnaeus                            17
      Crayfish, Orconectes sp., Cambarus sp., Procambarus     22
                 sp., or Pacifastacus leniusculus             22
      Stoneflies, Pteronarcys sp.                             12
      Mayflies, Baetis sp. or Ephemerella sp.                 17
                Hexagenia limbata or tl. bilineata             22
      Midges, Chironomus sp.                                  22
Marine and estuarine
   Vertebrates
      Sheepshead minnow, Cyprinodon variegatus                 22
      Mummichog, Fundulus heteroclitus                         22
      Longnose killifish, Fundulus similis                     22
      Silverside, Menidia sp.                                  22
      Threespine stickleback, Casterosteus aculeatus           22
      Pinfish, Lagodon rhomboides                              22
      Spot, Leiostomus xanthurus                               22
      Shiner perch, Cymatogaster aggregata                     12
      Pacific staghorn sculpin, Leptocottus  armatus            12
      Sanddab, Citharichthys stigmaeus                         12
      Flounder, Paralichthys dentatus, P_. lethostigma          22
      English sole, Parophrys vetulus                          12

-------
TABLE 1.  (Cont'd)
                                                                     o
Species                                         Test Temperature (°C)
Marine and estuarine
                3.
   Invertebrates
      Shrimp, Penaeus setiferus, P_. duorarum, or             22
         P_. aztecus
      Grass shrimp, Palaemonetes sp.                         22
      Shrimp, Crangon sp.                                    22
      Oceanic shrimp, Pandalus jordani                       12
      Blue crab, Callinectes sapidus                         22
      Dungeness crab, Cancer magister                        12
      Mysid shrimp, Mysidopsis sp., Neomysis sp.             22

Q
 Freshwater amphipods, daphnids, and midge larvae and shrimp should be
cultured and tested at the recommended test temperature.  Other in-
vertebrates should be held and tested within 5°C of the temperature of
the water from which they were obtained.  They should be tested at the
recommended test temperature if it is within this range; otherwise, they
should be tested at the temperature from the series 7, 12, 17, 22, and
27°C that is closest to the recommended test temperature and is within
the allowed range.

-------
HOLDING AND HANDLING

     Disinfect holding tanks or chambers with 0.5 percent commercial
bleach for one hour. Brush thoroughly with the disinfectant.  Rinsed well
between groups of organisms.  Other equipment used to handle organisms
must be disinfected with 0.5 percent commercial bleach, or 30-percent
formalin.

     When feasible, tanks must receive an uncontaminated, constant-quality
water in a flow-through system with a flow rate of at least 2 tank-volumes
per day.  Otherwise, a recirculation system where the water flows through
a charcoal filter to remove dissolved metabolites or passes by an ultra-
violet light for disinfection is necessary.  Only as a last resort should
a dechlorinated tap water be used for freshwater organisms or a synthetic
salt water for marine organisms.

     When organisms are first brought into the facility, they must be
quarantined for a minimum of 10 days.  To avoid unnecessary stress after
collection and transportation, organisms should not be subjected to more
than a 3°C change in water temperature or 3-ppt change in salinity in any
12-hour period.  Invertebrates should be held within 5°C of the temperature
of the water from which they were obtained.  To maintain organisms in good
condition during holding, crowding should be avoided.  Dissolved oxygen
must be greater than 40 percent saturation for warm water species and
greater than 60 percent saturation for cold water species.  Aeration may
be used if necessary.  Organisms should be fed at least once a day, and
excess food and fecal material should be removed from the bottom of the
tanks at least twice a week by siphoning.  Organisms should be observed care-
fully for signs of disease, stress, physical damage, and mortality.  Dead and
abnormal specimens should be removed as soon as observed.  A daily log of
feeding, behavioral observations, and mortality must be maintained.

     Organisms should be handled as little as possible.  When handling is
necessary, it should be done as gently, carefully, and quickly as possible
to minimize stress.  Organisms that touch dry surfaces or are dropped or
injured during handling must be discarded. Dipnets are best for handling
larger organisms.  Such nets are commercially available or can be
made from small-mesh nylon netting, nylon or silk bolting cloth, plankton
netting, or similar material.  Nets coated with urethane resin are best
for handling catfish.  Smooth glass tubes with rubber bulbs should be used
for transferring smaller organisms such as daphnids and midge larvae.

DISEASE TREATMENT

     During holding, fresh and salt water fish should be chemically treated
to cure or prevent disease as recommended in Table 2.  However, if the fish
are severely diseased, it is advisable to discard the entire lot.  When
invertebrates become diseased, they should be discarded.  Tanks which are
contaminated with disease-causing microorganisms must be disinfected with
0.5 percent commercial bleach.
                                     10

-------
    TABLE 2.  RECOMMENDED PROPHYLACTIC AND THERAPEUTIC TREATMENTS FOR
                             FRESHWATER FISH3
Disease
  Chemical
Concentration  Duration of
   (mg/1)        Treatment
External      Oxytetracycline hydrochloride
Bacteria         (water soluble)

              Procaine Penicillin G in
              Dihydrostreptomycin sulfate
              solution (Franklin Lab)

              Benzalkonium chloride
              (HYAMLNE 1622 )

              Nitrofurazone (water mix)

              Neomycin sulfate
                                    25
                 30-60 min
Monogenetic
trematodes
fungi, and
external,
        d
protozoa
Formalin plus zinc-free
malachite green oxalate

Formalin

Potassium permanganate

Sodium chloride


DEXONR (35% Active Ingredient)
                               (3ml/100 gal)    48-72 hrs
     1-2


     3-5b

     25


     25
      0.1

    150-250

      2-6

  15000-30000
   2000-4000

      20
 30-60 min


 30-60 min

 30-60 min


  1-2 hrs


 30-60 min

 30-60 min

5-10 min dip
     (e)

 30-60 min
Parasitic
copepods
Trichlorf on
(MASOTEN )
0.25 Continuous
 This table indicates the order of preference of treatments that have been
reported to be effective,but their efficacy against diseases and toxicity
to fish may be altered by temperature or water quality.  Caution; test
treatments on small lots of fish before making large-scale applications.
Fish should not be treated the first day they are in the facility.
                                      11

-------
TABLE 2 (continued).

    Before using a treatment other than those listed in this table, addi-
    tional information should be obtained from sources such as Davis (1953),
    Hoffman and Meyer (1974), Reichenbach-Klinke and Elkan (1965), Snieszko
    (1970), and van Duijn (1973).

b.  Active ingredient.

c.  Treatment may be accomplished by (1) transferring the fish to a static
    treatment tank and back to a holding tank; (2) temporarily stopping
    the flow in a flow-through system, treating the fish in a static
    manner, and then resuming the flow to flush out the chemical; or (3)
    continuously adding a stock solution of the chemical to a flow-through
    system by means of a metered flow or the technique of Brungs and Mount
    (1967).

d.  One treatment is usually sufficient except for Ichthyophthirius ("Ich"),
    which must be treated daily or every other day until no sign of the
    protozoan remains.  This may take 4-5 weeks at 5-10°C and 11-13 days
    at 15-21°C.  A temperature of 32°C is lethal to "Ich" in one week.

e.  Minimum of 24 hours, but may be continued indefinitely.

f.  Continuous treatment should be employed in static or flow-through
    systems until no copepods remain, except that treatment should not be
    continued for more than 4 weeks and should not be used above 27°C.
                                       12

-------
TRANSPORTATION AND ACCLIMATION

     Organisms reared and held at a fixed facility are transported to the
test site in the water in which they were reared and held.  If the labora-
tory is mobile, the acclimation tank can be used in transporting organisms
from the rearing and holding facilities to the test site. At the test site,
dilution water (receiving water) is pumped to the laboratory for use in the
acclimation of the organisms.  If dilution water is not readily accessible,
it can be transported to the laboratory and stored in a tank for use in
the acclimation procedure and toxicity test. During transport and acclimation
the change in water temperature must not exceed 3°C within 12 hours, and the
concentration of dissolved oxygen must not fall below 40 percent of saturation
for warm water species and 60 percent of saturation for cold water species.

     Upon arriving at the test site, the organisms are acclimated to the test
dilution water and temperature by gradually changing from 100-percent holding
water to 100-percent dilution water over a period of 24 hours.  All organisms
must be exposed to 100-percent dilution water for at least 24 hours before
they are used for the tests, and must be held at the test temperature (±2°C)
for at least 24 hours before tests are begun.

     A group of organisms must not be used for a test if they appear to be
diseased or otherwise stressed.  They should not be used if more than
5 percent die during the 48 hours immediately preceding the test.  If a group
of organisms fail to meet these criteria, all organisms must be discarded
and a new group obtained.  The same acclimation procedure must be followed
for the new group of organisms.

    Marine organisms, at times, may require acclimation to a series of
salinities ranging from 5 to 35 ppt.  The reason for such ranges is that
most effluents discharged into the marine environment consist of adulter-
ated freshwater.   Therefore, when diluted with the salt water, the higher-
percent effluent volumes will usually have a low salinity.  In these cases,
the salinity will be inversely proportional to the percent effluent
volume.
                                     13

-------
                                SECTION 4

                             DILUTION WATER
     A dilution water is acceptable if healthy test organisms survive
in it through the acclimation period and toxicity test without showing signs
of stress, such as discoloration or unusual behavior.   For efEluent toxi-
city testing, the dilution water should be a representative sample of the
receiving water, and should be obtained from a point as close as possible
to, but upstream of or outside of,  the zone influenced by the effluent.
It is preferable to pump the dilution water continuously to the acclima-
tion tank and dilutor.  However, it may be more practical to transport
batches of water in tanks to the testing site as the need arises, and
then continuously pump water to these systems.

     In an estuarine environment, the investigator should collect uncon-
taminated water having a salinity as near as possible to the salinity of
the water at the receiving site.

     Pretreatment of the dilution water should be limited to filtration
through a nylon sieve having 2-mm or larger holes to remove debris
and/or break up large floating or suspended solids.  The water should be
obtained from the receiving water as close as possible to the time the
test begins.  It should not be obtained more than 96 hours prior to
testing.  If acceptable dilution water cannot be obtained from the
receiving water, some other uncontaminated, well-aerated surface or
ground water, or commercially available media can be used.  This water
must have a total hardness, total alkalinity, and specific conductance
within 25 percent, and pH within 0.2 units, of the receiving water at
the time of testing.

     If  the substitute dilution water must be modified, reconstituted
water must be prepared for use as the diluent.  Recommended procedures
are given in Tables 3 and 4.  There are also commercially available salt
water media such as INSTANT OCEANR and RILA SALTSR.

     With highly toxic effluents requiring very large volumes of dilu-
tion water, it may be convenient to locate the testing facility near the
source of the dilution water, and transport the effluent.
                                      14

-------
          TABLE 3.   PREPARATION OF RECONSTITUTED FRESH WATERS

3a.  Quantities of (mg/1) reagent grade chemicals required to prepare
     recommended reconstituted fresh waters and the resulting water
     qualities.
Water
Type
Reagent Added
NaHCO. CaSO. .2H00 MgSO. KCL
3 42 4
Final Water Quality
, Alka-
b c c
pH Hardness linity
Very
Soft
Hard
Very
soft


hard
12
48
192
384
7.
30.
120.
240.
5
0
0
0
7.
30.
120.
240.
5
0
0
0
0.5
2.0
8.0
16.0
6.
7.
7.
8.
4-6.8
2-7.6
6-8.0
0-8.4
10-13
40-48
160-180
280-320
10-13
30-35
110-120
225-245
3b.  Quantities of reagent-grade chemicals to be added to aerated, soft
     reconstituted freshwater for buffering pH.   The solutions should not
     be aerated after addition of these chemicals.
                  Volume (ml) of solution added to 15 liters of water
                   l.ON NaOH
1.0 m KH0PO.
        2  4
0.5 m
6.0
6.5
7.0
7 c
/ . J
8.0
8.5
9.0
9.5
10.0
1.3
5.0
19.0


19.0
6.5
8.8
11.0
16.0
80.0
30.0
30.0


20.0
	
	
	
____
	
	
	


	
40.0
30.0
20.0
18.0
a. From Marking and Dawson (1973).
b. Approximate equilibrium pH after aeration and with fish in water.
c. Expressed in mg/1 as CaCO .
d. Approximate equilibrium pH with fish in water.
                                      15

-------
       TABLE 4. PREPARATION OF RECONSTITUTED SEA WATERa'b

Add the following reagent-grade chemicals in the amounts and order listed

to 890 ml water.  Each chemical must be dissolved before another is added.
       Chemical                                        Amount
                                                          3 mg

          SrCl2'6H20                                     20 mg

          H3B03                                          30 mg

          KBr                                           100 mg

          KC1                                           700 mg
          CaCl2-2H20                                    1.47 g

          Na2S04                                        4.00 g

          MgCl2'6H20                                   10.78 g

          NaCl                                         23.50 g
          Na2Si03'9H20                                   20 mg
          Na4EDTA(c)                                      1 mg
          NaHCO                                         200 mg
a.  If the resulting solution  is diluted  to  1  liter,  the  salinity should be
    34±0.5 g/kg, and the pH  8.0±0.2.  The  desired  salinity is  attained by
    dilution at  time of use.

b.  From Kester  et^ a±.  (1967),  Zaroogian  et^  al.  (1969), and Zillioux
    et al.  (1973).

c.  Tetrasodium  ethylenediaminetetraacetate. This should  be omitted when
    toxicity tests are conducted with metals.  When tests  are  conducted
    with plankton  or larvae, the EDTA should be omitted and the medium
    should be  stripped of  trace metals  (Davey  et^  al_. ,  1970).
                                      16

-------
                                  SECTION 5

                        EFFLUENT SAMPLING AND HOLDING
SAMPLING
     The effluent sampling point must be the same as that specified in the
NPDES discharge permit.   Conditions for exception would be:  (1) better
access to a sampling point between the final treatment and the discharge
outfall, or (2) if the processed waste is chlorinated prior to discharge
to the receiving waters, the sampling point may be located prior to con-
tact with the chlorine if the purpose of the test is to determine toxicity
levels of the unchlorinated effluent.  Sampling should be based on an under-
standing of the short and long-term operations and schedules of the dis-
charger.  It is desirable to evaluate an effluent sample that most closely
represents the "normal" or "typical" discharge and operating conditions
of the plant in question.  The retention time of the effluent in the waste
water treatment facility, as indicated in Par. A.l.a-c and B.l-3 below,
must be measured using dye studies.  The only way in which the sample may
be altered prior to testing is by filtering through a TEFLON  or stainless
steel screen with 2-mm or larger holes.

     A.  Flow-through test

         1.  If the industrial or municipal facility discharges continu-
             ously, the effluent should be pumped directly and continuously
             from the discharge line to the dilutor system for the
             duration of the test.  The use of the effluent grab samples
             should be avoided.  However, if the effluent cannot be pumped
             directly and continuously to the dilutor system, the fol-
             lowing alternative methods may be employed for collection
             of the effluent:

             a.  When the measured minimum retention time of the effluent
                 is less than 96 hours, as determined from dye studies, a
                 6—hour flow—proportional composite sample must be col-
                 lected and transported to the dilutor every 6 hours for
                 the duration of the test.

             b.  When the measured minimum retention time of the effluent
                 is between 4 days (96 hours) and 14 days, as determined
                 from dye studies, then a 24-hour flow-proportional com-
                 posite sample may be collected daily for the duration
                 of the test.
                                      17

-------
              c.  When the measured minimum retention time of the effluent
                  is greater than 14 days, as determined from dye studies,
                  a single grab sample may be collected daily for the
                  duration of the test.

          2.  If the industrial or municipal facility discharges inter-
              mittently (i.e., less than 96 hours of continuous flow),
              a flow-proportional composite sample may be taken for
              an 8-hour operating period or for the duration of the plant
              operating schedule.

      B.  Static test

          If a flow-through test cannot be used, a static test may be
          conducted with effluent collected by one of the following methods:

          1.  When the measured minimum retention time of the effluent is
              less than 96 hours, as determined with dye studies, a flow
              proportional composite sampler may be employed to collect the
              samples.  Four consecutive 6-hour flow-proportional composite
              samples may be collected and used in setting up 4 separate
              static tests.

          2.  When the measured minimum retention time of the effluent is
              between 4 days (96 hours) and 14 days, as determined with dye
              studies, a 24-hour flow-proportional composite sample may be
              used in the test.

          3.  When the measured minimum retention time of the effluent is
              greater than 14 days, as determined with dye studies, a
              single grab sample may be collected and used in the test.

HOLDING

     Effluent grab samples must be stored in covered, unsealed containers.
Violent agitation must be avoided.  However, undissolved materials must be
uniformly dispersed by gentle agitation.  This agitation must immediately
precede adjustment of any aliquot of the effluent to test temperatures
before adding it to the dilution water.  Although it is desirable to
refrigerate samples prior to the test, it is often convenient to store
samples in a constant-temperature water bath or controlled-environment
room at the temperature at which the test is conducted.  The test should
be initiated as soon as possible, but no later than 8 hours after collection
of the effluent.

     The persistence of the toxicity of an effluent may be a factor in deter-
mining specific toxicity limits in an NPDES permit, and is determined by
measuring its toxicity upon collection and again after holding 96 hours.
If after holding the effluent 96 hours, its toxicity has not decreased 50%
or more, it is classified as persistent.  (When special tests, such as
persistence are conducted, the exact methodology must be detailed in the
report.)

                                     18

-------
                                SECTION 6

                             TEST PROCEDURE

RANGE-FINDING TEST

     Unless the approximate toxicity of the effluent is already known, it
is necessary to conduct an abbreviated, preliminary, range-finding test to
determine the concentrations that should be used in the definitive tests.
This test can be either a static or flow-through test.  However, the test
most often used is an abbreviated static test in which groups of 5
organisms are exposed to three to five widely-spaced effluent dilutions,
and a control, for 8 to 24 hours.

     Because the characteristics of the effluent and the receiving water
may vary significantly within short periods of time, the toxicity observed
in a range-finding test may not be representative of the toxicity of the
effluent.  If the range-finding test is to be conducted with the same
sample of the effluent with which the definitive test is to be conducted,
the duration of the range-finding test cannot exceed 8 hours (see limits
in holding time for effluents, p. 18).

DEFINITIVE TEST

Test Conditions-

     The determination of a LC50 or EC50 must employ a control and at
least five concentrations of effluent in an exponential series.  To
calculate the LC50 or EC50 with reasonable accuracy, a definitive test
must meet both of the following criteria:

     A.  Each concentration of the effluent must be at least 50
         percent of the preceding concentration.

     B.  One concentration must have killed  (or affected) more than 65
         percent of the organisms exposed to it, and one concentration other
         than the control must have killed (or affected) less than 35
         percent of the organisms.

     If 100-percent effluent does not kill (or affect) more than 65 percent
of the organisms exposed to it, the percentage of organisms killed (or
affected) by various levels of the effluent  in the receiving water must be
reported.

    The control shall consist of the same dilution water, conditions,
procedures, and organisms used in testing the effluent.  A test is not
acceptable if more than ten percent of the organisms die in the control.

Number of Test Organisms

     At least 20 organisms of a given species must be exposed to each  treat-
ment.  More than one species may be used in  the same test chamber in a

                                      19

-------
given test, if segregated.  One-half of the organisms of each species
exposed to each treatment should be placed in separate test chambers to
serve as replicates.  To qualify as true replicates, no water connections
can exist between replicate test chambers.  Randomization of treatments is
desirable.

     Test animals are normally captured for transfer from acclimation tanks
to test chambers by dip netting.  No more than 20 percent of the total
number of organisms transferred to each chamber should be added from a
given net capture.

Loading of Test Organisms

     For all tests, a limit must be placed on the weight of organisms
per liter of test solution.  This will minimize the depletion of dissolved
oxygen, the metabolic conversion of effluent constituents, the accumu-
lation of metabolic waste products, and/or stress induced by crowding, any
of which could significantly affect the test results.

     For flow-through tests, loading in the test chambers must not exceed
5 grams per liter at temperatures of 20°C or less, or 2.5 grams per liter
at temperatures above 20°C.

     For static tests, loading in the test chambers must not exceed 0.8
grams per liter at temperatures of 20°C or less and 0.4 grams per liter at
temperatures above 20°C.

Test Temperature

     For flow-through tests, it is desirable to hold the temperature
within ± 2.0°C of the acclimation temperature throughout the test.  This
can be accomplished by passing the effluent and/or dilution water through
separate stainless steel coils immersed in a heating or cooling water bath
prior to entering the dilutor system.

     For static tests, the temperature may be that at which the test
organisms were held prior to transportation or acclimation at the site.
The instantaneous ambient temperature should not vary more than ±2°C at any
time during the test.

Oxygen Requirements and Aeration

     Aeration may alter the results of toxicity tests and, as a general
rule, should not be employed. It can reduce the apparent toxicity of an
effluent by stripping it of highly volatile toxic substances, or increase
its toxicity by altering the pH. However, the dissolved oxygen (DO) in the
test solution should not be permitted to fall below 40 percent saturation
for warm water species and 60 percent saturation for cold water species.
In most flow-through tests, DO depletion is not a problem in the test
chambers because aeration occurs as the liquids pass through the dilutor
system.
                                      20

-------
     If the DO concentration decreases to a level that would be a
source of additional stress, the turnover rate of the solutions in the
test chambers must be increased sufficiently to maintain acceptable DO
levels. If the increased turnover rate does not maintain adequate DO
levels, aerate the dilution water prior to the addition of the effluent,
and aerate all test solutions.

     Caution must be exercised to avoid excessive aeration, and it
should be used only as a last resort in maintaining adequate DO levels.
When aeration is used, the exact methodology must be detailed in the report.

Beginning the Test

     The test begins when the test organisms are first exposed to the
effluent.

     A.  Flow-through test

         The dilutor system should be in operation 24 hours prior to
         the addition of the test organisms and the beginning of the test.
         During this period, necessary adjustments can be made in the
         percent effluent volumes, temperature, and flow rate through
         the test chambers.

     B.  Static test

         The effluent is added to the dilution water and mixed well
         by stirring with a glass rod. The test organisms are placed
         in the chambers within 30 minutes. This procedure conserves DO
         and is sufficient for the effluent to become evenly dispersed
         in the dilution water.

Feeding

     Organisms should not be fed during the test.  However, feeding is per-
mitted when using newly hatched or very young organisms.  Feeding in
a  flow-through system minimizes problems encountered in a static system,
such as alteration of test dilution and build-up of food and metabolic
waste products, resulting in a greater oxygen demand.

Duration

     The test duration may range from a minimum of 8 hours to 96 hours,
depending on whether it is a range-finding test or a definitive  test and
the purpose of the test.
                                      21

-------
                                SECTION 7

                               TEST RESULTS

BIOLOGICAL DATA

     The lengths and weights of the test organisms should be determined by
sacrificing and measuring representative organisms before the test or
by obtaining the lengths and weights of all surviving organisms at the
end of the test.  The number of dead (or affected) organisms in each
test container should be counted 24, 48, 72, and 96 hours after the
beginning of the test.  (See data sheet in Appendix, Fig. 1, p. 33).
Dead organisms must be removed at least once every 24 hours.

     Death is the indicator most frequently used for studying toxicity
to aquatic organisms.  The criteria for death are usually no movement
(especially no gill movement in fish) and no reaction to gentle prod-
ding.  Death is not easily determined for some invertebrates.
The effect generally used for determining toxicity to daphnids and midge
larvae is immobilization, which is defined as the inability to move
except for minor activity of appendages, and with crabs, crayfish, and
shrimp the effects used are immobilization and loss of equilibrium.
Other effects can be used for determining an EC50, but the effect and
its definition must always be reported.  General observations on such
things as erratic swimming, loss of reflex, discoloration, changes in
behavior, excessive mucus production, hyper-ventilation, opaque eyes,
curved spine, hemorrhaging, molting, and cannibalism should be reported.

CHEMICAL AND PHYSICAL DATA

     The dissolved oxygen concentration and pH must be measured at the
beginning of the test, and every 24 hours thereafter, in the control
and in the high, medium, and low effluent concentrations, for the
duration of the test. The specific conductance, total alkalinity,
total hardness, and salinity, where applicable, should be measured at
the beginning of the test in the control and the high, medium and low
effluent concentrations.  There may be a build-up of ammonia in the static
toxicity tests.  It is advisable, therefore, to measure  the concentration
of total ammonia nitrogen in the control, high, medium,  and low effluent
concentrations at the beginning and end of each static test.  The percentage
of un-ionized ammonia in the test containers can be determined from
Tables 5 and 6.  Temperature should be recorded at  least hourly in at least
one container during the acclimation period and test.
                                      22

-------
     TABLE 5.  PERCENTAGE OF AMMONIA THAT IS UN-IONIZED IN DISTILLED
                   WATER AT VARIOUS TEMPERATURES AND pH'Sa
                        PERCENT UN-IONIZED AMMONIA
Temperature
CO
pH
6.0 6.5 7.0 7.5 8.0 8.5 9.0 9.5
10.0
     7

    12

    17

    22

    27
0.01   0.05   0.15   0.46   1.45   4.44   12.8   31.7   59.5

0.02   0.07   0.22   0.68   2.13   6.44   17.9   40.8   68.5

0.03   0.10   0.32   1.00   3.08   9.14   24.1   50.1   76.1

0.04   0.14   0.45   1.43   4.39  12.7    31.5   59.2   82.1

0.06   0.21   0.65   2.03   6.15  17.2    39.6   67.4   86.8
                  TABLE 6.  SALINITY CORRECTION FACTOR
             Salinity (g/kg)
                                    Factor
                   5

                  10

                  15

                  20

                  25

                  30

                  35
                                     0.82

                                     0.80

                                     0.77

                                     0.75

                                     0.73

                                     0.72

                                     0.72
a. Skarheim (1973), and Thurston et al.(1974).
b. These values can be corrected for salinity by multiplying by the approp-
   riate factor from Table 6.
c. The resulting values should be within about ±5% over the range of pH
   from 6 to 9, and temperature from 0° to 30°C.
                                      23

-------
     Water samples collected for chemical analysis should be taken
midway between the top, bottom, and sides of the test containers
and should not include any surface scum or material stirred up from
the bottom or sides.

     Methods used for chemical analysis must be those specified in
Section 304(g) of the Federal Water Pollution Control Act
Amendments of 1972 (Chemical Methods Manual, USEPA, 1977).

CALCULATION OF LC50 AND EC50

     For each set of data, the 96-hr LC50 or EC50 and its 95-percent
confidence limits must be calculated on the basis of the initial volume
percent of the effluent in the test solutions.  The "volume percent"
equals "(100 x volume of effluent)/(volume of effluent + volume of
dilution water)."  If other (24-,48-,72-hr) LC and EC values are
calculated, their 95 percent confidence limits must also be determined.
A variety of methods are available to calculate a LC and EC (Finney,
1964, 1971). The most widely used are the log-concentration versus-
percent-survival, probit, logit, moving-average, and Litchfield-Wilcoxon
(1949) methods.

     Two examples of calculating a  LC50 using a hypothetical set of
data have been provided in the Appendix.  The two methods used in the
calculations are:   (1) the Litchfield-Wilcoxon and, (2) the log-
concentration-versus-percent-survival.  If more than  10 percent of the
control organisms die, none of the previously mentioned methods may
be used in calculating LC and EC values, and the remaining test
results must be used with caution in evaluating the toxic effect.

REPORTS

A report of the results of a test must include the following:

A.  The name of the test method, investigator and laboratory, and
    the date the test was conducted.

B.  A detailed description of the effluent, including its source,
    date and time of collection, composition, known physical and
    chemical properties, and variability.

C.  The source of the dilution water, the date and time of its
    collection, its chemical characteristics, and a description of
    any pretreatment.

D.  Detailed information about the test organisms, including scientific
    name,  length and weight, age, life stage, source, history, observed
    diseases,  treatments, and acclimation procedure used.
                                      24

-------
E.  A description of the test procedure: the test chambers, including
    the depth and volume of solution; the way the test was begun; the
    number of organisms per treatment; and the loading.  For the flow-
    through system, the water volume changes per 24 hours in each test
    chamber must be calculated and reported.

F.  The definition of the adverse effect (death, immobility, etc.) used
    in the test, and a summary of general observations on other effects or
    symptoms.

G.  The number and percentage of organisms in each test chamber (including
    the control chambers) that died or showed the "effect" used to measure
    the toxicity of the effluent.

H.  A 24-, 48-, 72-, and 96-hr LC50 or EC50 value for the test
    organisms, depending on the duration of exposure. If 100 percent
    effluent did not kill or affect more than 65 percent of the test
    organisms, report the percentage of the test organisms killed or
    affected by various concentrations of the effluent.

I.  The 95-percent confidence limits for the LC50 and EC50 values and the
    method used to calculate them.

J.  The methods used for and the results of all chemical analyses.

K.  The average and range of the acclimation temperature and the test
    temperature.

L.  Any deviation from this method.

M.  Any other relevant information.
                                     25

-------
                               REFERENCES
 1.   Abram,  F.S.H.  1973.   Apparatus for control of poison concentration
     in toxicity studies  with fish.   Water Res.  7:1875-1879.

 2.   Bengtsson,  B.E.  1972. A simple principle for dosing apparatus in
     aquatic systems.   Arch.  Hydrobiol.  70:413-415.

 3.   Brungs, W.  A.,  and D. I. Mount. 1967. A device for continuous treatment
     of fish in  holding chambers.  Trans.  Amer.  Fish.  Soc.  96:55-57.

 4.   Cline,  T.F. and G. Post.  1972.  Therapy for trout eggs  infected with
     Saprolegnia.   Prog.  Fish-Cult.   34:148-151.

 5.   Davey,  E.W. ,  J.H.  Gentile,  S.J. Erickson,  and P.  Betzer.,  1970.
     Removal of  trace metals from marine culture media.  Limnol.  Oceanogr.
     15:486-488.

 6.   Davis,  H.S. 1953.  Culture and diseases of  game fishes. Univ.  Calif.
     Press,  Berkeley.  332 pp.

 7.   DeFoe,  D.L. 1975.  Multichannel toxicant injection system for flow-
     through bioassays.  J. Fish.  Res.  Bd. Can.  32:544-546.

 8.   Finney, D.J.  1964.  Statistical Method in  Biological Assay,  2nd ed.
     Hafner  Publ.  Company, New York.  668 pp.

 9.   Finney, D.J.   1971.   Probit Analysis.  Cambridge Univ. Press,
     London.  333 pp.

10.   Freeman, R. A.  1971.  A constant flow delivery device for chronic
     bioassay.  Trans.  Amer. Fish. Soc.  100:135-136.

11.   Granmo, A.  and S.  0. Kollberg.  1972.  A new simple water flow system
     for accurate continuous flow tests.  Water Res.   6:1597-1599.

12.   Hoffman, G.L.,  and F. P. Meyer. 1974. Parasites of freshwater fishes.
     THF Publ.,  Inc.,  Neptune City, N.J. 224 pp.

13.   Kester, D.R., I.W. Dredall, D.N. Connors,  and R.M. Pytokowicz.   1967.
     Preparation of artificial seawater.  Limnol. Oceanogr.  12:176-179.

14.   Lichatowich, J.A., P.W. O'Keefe, J.A. Strand, and W. L.  Templeton.
     1973.  Development of methodology and apparatus for the  bioassay of  oil.
     In:  Proceeding of Joint Conference on Prevention and Control of Oil
     Spills.  American Petroleum Institute, U.S.  Environmental Protection
     Agency and U.S. Coast Guard, Washington, D.C.  pp. 659-666.

15.   Litchfield, J.T. Jr. and F. Wilcoxon.  1949.  A simplified method of
     evaluating dose-effect experiments.  J. Pharm. Exp. Ther.  96:99-113.
                                     26

-------
16.  Lowe, J.I.  1964.   Chronic exposure of spot, Leiostomus xanthurus, to
     sublethal concentrations of toxaphene in seawater.  Trans. Amer. Fish.
     Soc.  93: 396-399.

17.  Marking, L. L..,  and V.  K. Dawson. 1973. Toxicity of quinaldine
     sulfate to fish.  Invest. Fish Contr. No. 48., U.S. Fish & Wildlife
     Service, Washington, B.C. 8 pp.

18.  Mount, D. I. and  W. A. Brungs.   1967.  A simplified dosing apparatus
     for fish toxicological studies.  Water Res.  1:21-29.

19.  Nebeker, A. V. and A. E. Lemke.  1968.  Preliminary studies on the
     tolerance to aquatic insects to heated waters.  J. Kans. Entomol.
     Soc.  41:413-418.

20.  Reichenbach-Klinke, H.,  and E.  Elkan. 1965. The principal diseases of
     lower vertebrates.  Academic Press, New York, 600 pp.

21.  Riley, C. W.  1975.  Proportional diluter for effluent bioassays.
     JWPCF.  47:2620-2626.

22.  Schimmel, S. C.,  D. J. Hansen,  and J. Forester.  1974.  Effects of
     Aroclor 1254 on laboratory-reared embryos and fry of sheepshead min-
     nows (Cyprinodon  variegatus).  Trans. Amer. Fish. Soc.  103:582-586.

23.  Schimmel, S. C. and D. J. Hansen.  1974.  Sheepshead Minnow (Cyprinodon
     variegatus):  An  estuarine fish suitable for chronic (entire life-cycle)
     bioassays.   Proceedings of the 28th Annual Conference of the South-
     eastern Association of Game and Fish Commissioners.  pp. 392-398.

24.  Shumway, D. L. and J. R. Palensky.  1973.  Impairment of the flavor of
     fish by water pollutants.  Ecological Research Series No.  EPA-R3-73-
     010.  U. S. Environmental Protection Agency, Washington, D. C.  80 pp.

25.  Skarheim H. P.  1973.  Tables of the fraction of ammonia in the
     undissociated form.  SERL Report No. 73.-5.  University of California,
     Berkeley, 33 pp.

26.  Sniewzko, S. F. (ed.). 1970. A symposium on diseases of fishes and
     shellfishes. Spec.  Publ. No. 5, Amer. Fish. Soc., Washington,  D.C.
     526 pp.

27.  Sprague, J. B. 1969.  Measurement of pollutant toxicity to fish.  I.
     Bioassay methods  for acute toxicity.  Water Res.  3:793-821.

28.  Thurston, R.V., R.  C. Russo, and K. Emerson.  1974.  Aqueous ammonia
     equilibrium calculations.  Tech. Rep. No. 74-1.  Fisheries Bioassay
     Laboratory, Montana State University, Bozeman.  18 pp.

29.  U.S. Environmental Protection Agency. 1972. Recommended bioassay
     procedure for fathead minnow Pimephales promelas Rafinesque chronic
     tests. U.S. Environmental Protection Agency, National Water Quality
     Laboratory, Duluth, Mn.  13 pp.

                                     27

-------
30.  U.S. Environmental Protection Agency.  1974.  Methods for chemical
     analysis of water and wastes. U.S. Environmental Protection Agency,
     National Environmental Research Center, Methods Development and
     Quality Assurance Research Laboratory, Cincinnati, Oh. 298 pp.

31.  U.S. Environmental Protection Agency.  1975.  Methods for acute
     toxicity tests with fish, macroinvertebrates and amphibians.
     U.S. Environmental Protection Agency,  National Environmental
     Research Center, National Water Quality Research Laboratory,
     Duluth, Mn. 61 pp.

32.  van Duijn, C., Jr. 1973. Diseases of fishes. 3rd ed., Charles C.
     Thomas Publ., Springfield, II. 309 pp.

33.  Zaroogian, G. E., G. Pesch, and G. Morrison.  1969.  Formulation of an
     artificial sea water media suitable for oyster larvae development.
     Amer. Zool.  9:1141.

34.  Zillioux, E. J., H.R. Foulk, J.C. Prager, and J. A. Cardin.  1973.
     Using Artemia to assay oil dispersant toxicities.  JWPCF.  45:2389-
     2396.
                                      28

-------
                                  APPENDIX

A.  LITCHFIELD AND WILCOXON ABBREVIATED METHOD OF DETERMINING THE LC50

General Procedure

Step 1:  Tabulate the data (see sample data sheet, Fig. 1, p. 33) showing
the percent-effluent volumes used, the total number of organisms exposed to
each percent-effluent volume, the number of affected organisms, and the
observed percent-affected organisms (see Example 1 below).  Do not list
more than 2 consecutive 100 percent affects at the higher percent-effluent
volumes or more than two consecutive 0 percent affects at the lower percent-
effluent volumes.

Step 2:  Plot the percent-affected organisms against the percent-effluent
volume on 2 cycle, logarithmic probability paper (Fig. 2), except for
0 percent or 100 percent affect values.  With a straight edge, fit a
temporary line through the points, particularly those in the region of
40 percent to 60 percent affects.

Step 3:  Using the line drawn through the points, read and list an
"expected" percent affect for each percent-effluent volume tested.  Disre-
gard the "expected" percent value for any of the percent volumes less than
0.01 or greater than 99.99.  Using the expected-percent-affect, calculate
from Table 7 a "corrected" value for each 0 percent or 100 percent affect
obtained in the test.  (Since the expected values in the table are whole
numbers, it will be necessary to obtain intermediate values by interpolation.)
Plot these values on the logarithmic probability paper (Fig. 2) used in
Step 2 and inspect the fit of the line to the completely plotted data.  If
after plotting the corrected expected values for 0 percent and 100 percent
affected, the fit is obviously unsatisfactory, redraw the line and obtain
a new set of expected values.

Step 4:  List the difference between each observed (or corrected) value and
the corresponding expected value.  Using each difference and the corres-
ponding expected value, read and list the contributions to Chi-square (Chi2)
from Fig. 3 ( a straight edge connecting a value on the Expected-Percent
Affected scale with a value on the Observed-Minus-Expected scale, will
indicate at the point of intersection of the Chi^ scale, the contribution to
Chi2. Sum the contributions to Chi2 and multiply the total by the average
number of organisms per effluent volume, i.e., the number of organisms used
in K concentrations divided by K, where K is the number of percent-affected
organism values plotted. The product is the "calculated" Chi2 of the line.
The degrees of freedom (N) are 2 less than the number of points plotted,
i.e., N =K-2.  If the calculated Chi2 is less than the Chi2 given in Table 8
for N degrees of freedom, the data are non-heterogeneous and the line is
a good fit.  However, if the calculated Chi2 is greater than the Chi2 given
in Table 8 for N degrees of freedom, the data are heterogeneous and the line
is not a good fit.  In the event a line cannot be fitted (the calculated
Chi2 is greater than the tabular Chi2), the data can not be used to calcu-
late a LC50 or EC50.  Litchfield and Wilcoxon provided an alternate
method for calculating the 95 percent confidence limits under these cir-
cumstances.  However, the toxicity test should be repeated.
                                     29

-------
Step 5:  Determine the confidence limits of the LC50.

a.  Read from the fitted line (Fig. 2), the percent effluent volumes for
    the corresponding 16, 50, 84 percent affects (LC16, LC50 and LC84).
b.  Calculate the slope function, S, as:

                  S = LC84/LC50 + LC50/LC16
                                2

c.  From the tabulation of the data determine N', which is defined as the
    total number of test organisms used within the percent-affected-
    organism interval of 16 percent and 84 percent. Calculate the exponent
    (2.77//N1) for the slope function and the factor, f;LC50» used to
    establish the confidence limits for the LC50 (or EC50).

                            _  (2.77//N1)
                      £LC50   b

    ^ne ^LC50 can k£ obtained directly from the nomogram in Fig.4 by laying
    a straight-edge across the appropriate base and exponent values and
    reading the resultant "f" value.

e.  Calculate the confidence limits of the LC50 as follows:

    (1) Upper limit for 95% probability = LC50 X


    (2) Lower limit for 95% probability = LC50/fLC5()

Example

   Steps 1-4:  The data were tabulated and plotted (Fig. 2) and the
               expected values were read from the graph.

                   STEP ONESTEP THREESTEP FOUR
%
Effluent
Volume
3.2
5.6
10.0
18.0
32.0
56.0
100.0
Number of
Organisms
20
20
20
20
20
20
20
Number of
Affected
Organisms
0
1
11
7
12
18
20
Observed %
Affected
Organisms
0(.2)b
5
55
35
60
90
100 (99.0)
Expected %
(Fig. 2)
.6
3'5 a
(14.5)a
38.0
67.0
87.5
97.0
Observed
Minus
Expected
0.4
1.5
Aberrant
3.0
7.0.
2.5
2.0
Chi2
0.005
0.006
Value
0.004
0.024
0.006
0.014
                                                            Total    0.059
a. Percent-affected organisms at the 10 percent effluent volume  is obviously
   an aberrant value and should be omitted when fitting the line  in Step 2.

b. Step  3  "Corrected" affected values  from Table  7.
                                     30

-------
Step 4 (Cont.);
                  2
Calculation of Chi

a.  Total number of organisms used in 'K' concentrations = 120  = 20
                                                            6

b.  Calculated Chi2 = 20 x 0.059 = 1.18

c.  Degrees of Freedom (N)=K-2=6-2=4
                         2
d.  From Table 8, the Chi  for 4 degrees of freedom = 9.49

                      2                             2
e.  The calculated Chi  is less than the tabular Chi .  Therefore, it is
    assumed the line is a good fit, and the data are non-heterogeneous.

Step 5:

a.  From the fitted line in Fig. 2, determine the (percent) effluent
    concentrations corresponding to the 16%, 50% and 84% affected
    organism values:

b.  LC84 effluent concentrations  = 50.0%
    LC50     "          "         = 23.0%
    LC16     "          "           10.0%

c.  Calculate the slope function, 'S', as:

     S = LC84/LC50 + LC50/LC16 =  50.0/23.0 + 23.0/10.5
                   2                        2

       = 2.17 + 2.19  =  4.36  =  2.18
              2            2

d.  N' = 40  (From Figure 2)

e.  Calculate the exponent (N1) and factor, £,._,._

          = S2'77//N'= 2.182'77/1/4° = 2.182'77/6'32 - 2.18°'438 - 1.41  -
f.  Calculate the confidence limits of the LC50
    (1)  Upper limit for 95% probability = LC50 X fTr,_ = 23.0 X 1.4 = 32.2%
    (2)  Lower limit for 95% probability  = LCSO/f.^n = 23.0/1.4 = 16.4%
                                                  .LLOU
                                      31

-------
TABLE 7.  CORRECTED VALUES OF 0% OR 100% EFFECT

Expected
Value
Corrected Value

0
0
10 3.2
20 6.0
30 8.3
40 9.9
50
60 90.1
70 91.7
80 94.0
90 96.8

1
0.3
3.5
6.2
8.4
10.0
89.5
90.2
91.9
94.3
97.1

2
0.7
3.8
6.5
8.6
10.1
89.6
90.4
92.2
94.5
97.4

3
1.0
4.1
6.7
8.8
10.2
89.6
90.5
92.4
94.8
97.7

4
1.3
4.4
7.0
9.0
10.3
89.6
90.7
92.6
95.1
98.0

5
1.6
4.7
7.2
9.2
10.4
89.7
90.8
92.8
95.3
98.4


2
4
7
9
10
89
91
93
95
98

6
.0
.9
.4
.3
.4
.7
.0
.0
.6
.7

7
2.3
5.2
7.0
9.4
10.4
89.8
91.2
93.3
95.9
99.0

8
2.6
5.5
7.8
9.6
10.4
89.9
91.4
93.5
96.2
99.3

9
2.9
5.7
8.1
9.8
10.5
90.0
91.6
93.8
96.5
99.7
        TABLE 8.  VALUES OF Chi  (p = 0.05)
Degrees of Freedom (N)
1
2
3
4
5
6
7
8
9
10
Chi2
3.84
5.99
7.82
9.49
11.1
12.6
14.1
15.5
16.9
18.8
                          32

-------
                               FIG.  I.  DATA SHEET FOR EFFLUENT  TOXICITY TEST.
      Industry/Toxicant.
      Address	
      Contact	
      Effluent Serial Number .
      NPDES Permit Number
Beginning' Date.
Ending:   Date.
Test Organism	
.Time.
.Time.
Test Temperature Range.
Cone.
or
%
















Test
Container
Number
















Number of
Live Organisms
0
















24
















48
















72
















96
















Dissolved Oxygen
(mg/l)
0
















24
















48
















72
















96
















PH
0
















24
















48
















72
















96
















Total Alkalinity
(mg/l as CaCOs)
0
















24
















48
















72
















96
















Total Hardness
(mg/l as CaCO 3)
0
















24
















48
















72
















96

















0

















24

















48

















72

















96
















OJ
LO

-------
u>
-P-
                p

                ro
             O
           m z Zj
             a m
           ro   >
           -  CO ^

           y* 7J >
           •^ rn -
           > > >

           i°g

           H!3n
             O O
rn *-
                     So  o »
                                 % ORGANISMS AFFECTED

-------
 EXPECTED
    %
 AFFECTED
  50 =50
  70 —30

  80 —20
                                                        Chi
                                                          . 2
  90
    — 10
  95
  96
    	E
  97 -3
  98
     -1.0
 99.5

 99.6-4—4
 99.7
 99.8- -.2
 99.9-
99.95

99.96-

9 ft 97-
99.98	.02
     -.3
05

.04

.03
                      OBSERVED MINUS
                      EXPECTED

                          - 50

                          — 40

                          — 30


                          — 20
                                -10
                            5

                            4

                            3


                           - 2
                               - -.3
                                 .2
                               - -.OS
                                                      — 2.0
— 1.0
  .50

  .40

  .30


— .20
                                                           — .10
  .05
  .04

— .03


— .02
                                                           — .01
                                                        .005
                                                        .004

                                                        .003


                                                      — .002
                                                           —.001
       FIG. 3. NOMOGRAPH FOR OBTAINING Chi2 FROM
              EXPECTED  % AFFECTED AND
              OBSERVED-MINUS-EXPECTED (STEP 4).
                              35

-------
                   -p
                   '
EXPONENT
                    LC50
10 —
5.0 —
4X> —
3.0 —
-
2.0—
-
1.50 —
1.40—
1.30—
-
~
1.20—
	
_
-
1.10—
ta















m •+*+* •+*

- 100
- 50
— 10.0
— 5.0
— 4.0
— 3.0
-2.0
— 15
- 1.4
-1.3
— 1.2

— 1 10

- 1.05
— 1.04
- 1.03
- 1.02
— 4.0
-
— 3.0
-
— 2.0
— 1.5
-
— 1.0
— .90
— .80
— .70
— .60
— .50
-
— .40
-
— .30
-
_.20
FIG. 4. NOMOGRAPH FOR RAISING BASE S TO
      A FRACTIONAL EXPONENT
                   36

-------
B.  LOG-CONCENTRATION VERSUS PERCENT-SURVIVAL METHOD OF DETERMINING
    THE LC50

General Procedure

Step 1:  Plot the percent effluent volumes and the corresponding
percent survival on semi-logarithmic paper (Fig.  5).

Step 2:  Locate the 2 highest points on the graph which are separated
by the 50 percent survival line and connect them with a diagonal straight
line.  However, if one of the points is an aberrant value,  the next
lowest or highest percent-effluent volume is used.

Step 3:  Read on the scale for percent-effluent volume, the value of
the point where the diagonal line and the 50 percent survival line
intersect.  This value is the LC50 percent-effluent volume  for the
test.  If by chance one of the effluent concentrations happens to have
50 percent survival, no graphing is necessary.

Example

Step 1:  The percent-effluent volumes and the corresponding percent
survival data from the Litchfield and Wilcoxon example are  plotted in
Fig. 5.

Step 2:  The two highest points which are separated by the  50 percent
survival line (65 percent and 40 percent) are located and connected with a diagonal
diagonal straight line.  The percent survival in the 10 percent-effluent
volume was considered an aberrant value and, therefore, was omitted from the
evaluation.

Step 3:  An LC50 of 25.4 percent-effluent volume for the test was
derived from the point where the diagonal line and the 50 percent
survival line intersected in Fig.5.
                                     37

-------
    0   10  20  30  40  50  60   70  80   90  100
            % SURVIVAL
FIG. 5.  PLOTTED DATA AND FITTED LINE FOR
       LOG-CONCENTRATION VERSUS
       % SURVIVAL METHOD
               38

-------
C.  DILUTOR SYSTEMS, CONTROL PANEL AND EQUIPMENT LISTS

     Two proportional dilutor systems are illustrated:  (1) The Solenoid
Valve System, and (2) the Vacuum Siphon System.  The designs incorporate
features from devices developed by many other Federal and State pro-
grams, and have been shown to be very versatile for on-site bioassays
in mobile labs as well as in fixed (central) labs.  The Solenoid Valve
System is fully controlled by solenoids (Figs. 6, 7, & 8), and is the
preferred system.  The Vacuum Siphon System (Figs. 6, 9, & 10), how-
ever is acceptable.   Both systems employ the same control panel
(Fig. 14).

     If in the range-finding test, the LC50 of the effluent falls in the
concentration range, 5.6% to 100%, premixing is not required, and the
mixing chamber is by-passed by running a TEFLONR tube directly from
the effluent in-flow pipe to chamber E-l.  Chambers D-l and D-2 and the
mixing chamber are deactivated.

     To provide the capability of using the dilutor system to carry out
tests of the toxicity of pure compounds, the control panel is equipped
with an auxiliary power receptacle to operate a metering pump to deliver
an aliquot of the stock solution of the pure compound directly to the
mixing chamber during each cycle.  In this case, chamber D-l is
deactivated and chamber D-2 is calibrated to deliver a volume of
2000 ml, which is used to dilute the aliquot to the highest concen-
tration used in the toxicity test.
                                     39

-------
FIG. 6. PHOTOGRAPHS OF DILUTOR SYSTEMS:
      SOLENOID VALVE SYSTEM (LEFT), AND
      VACUUM SIPHON SYSTEM (RIGHT).
                   40

-------
          FLOW CONTROL
               VALVES-
                      NORMALLY OPEN
                      SOLENOID VALVES
 DILUTION WATER
  INFLOW
EFFLUENT
 INFLOW
   CYCLE
 COUNTER
  LAPSE
   TIME
   CLOCK
                                                      LIQUID LEVEL
                                                        SWITCH
                          DILUTION WATER CHAMBERS
                                MIXING CHAMBER
                                     NORMALLY OPEN
                                     SOLENOID VALVE
                   MAGNETIC
                    STIRRER
EFFLUENT
CHAMBERS













MIXING CHAMBERS
                      TEST CHAMBERS  4—20 LITERS CAPACITY

          FIG.  7. SOLENOID VALVE DILUTOR SYSTEM,
                  GENERAL DIAGRAM.
                                  41

-------
      n
         D-l
         n
                   D-2
 n
II
              D-3
n
ADJUSTABLE
STANOPIPE
DRAIN
                                        -DILUTION WATER CHAMBERS
                                     NORMALLY CLOSED
                                     SOLENOID VALVES
                                      6 mm 0 D DELIVERY TUBE


                                     NORMALLY  OPEN SOLENOID VALVE
                                      6 mm O.D. DELIVERY TUBE
                                             •EFFLUENT CHAMBERS
                                  ••	 NORMALLY CLOSED
                                      SOLENOID VALVE


                                  «	6 mm OD DELIVERY TUBE
                                      -MIXING CHAMBER
                                       IEOO ml CAPACITY
                                       10 mm OD DELIVERY TUBE
                                            -TEST  CHAMBER
                                             4-20 LITER CAPACITY
   FIG.  8. SOLENOID  VALVE DILUTOR SYSTEM,
            DETAILED DIAGRAM.
                                 42

-------
            for  dilution water  and
Solenoid System Equipment List

 1.  Diluter Glass.
 2.  Stainless Steel Solenoid Valves
     a.  3 - normally open, two-way, 55 psi, water, 1/4" pipe size,
         9/32" orifice size, ASCO 8262152, for incoming effluent and
         dilution water pipes and mixing chamber pipe.
     b.  1 - normally closed, two-way, 15 psi, water,  3/8" pipe size,
         3/8" orifice size, ASCO 8030B65, for D-2 chamber evacuation
         pipe.
     c.  12 - normally closed, two-way, 36 psi, water, 1/4" pipe
         size, 9/32" orifice size.  ASCO 8262C38, for  remaining dilution
         chamber and effluent chamber evacuation pipes.
 3.  Stainless steel tubing, seamless, 316-grade, austenetic.
     a.  10 ft - 3/8" OD, 0.035" wall thickness, for dilution water and
         effluent pipes.
     b.  60 ft - 1/4" OD, 0.035" wall thickness,
         effluent pipes.
     c.  1 ft - 3/4" OD, 0.035"-wall thickness, for standpipe in D-l
         chamber.
 4.  Swagelok tube connectors, stainless steel.
     a.  4 - male tube connectors, male pipe size 1/4", tube OD 3/8".
     b.  2 - male tube connectors, male pipe size 1/2", tube OD 3/8".
     c.  26 - male tube connectors, male pipe size 1/4, tube OD 1/4"
     d.  2 - male tube connectors, male pipe size 3/8", tube OD 3/8".
     e.  2 - male adapter, tube to pipe, male size 1/2", tube OD 3/8".
 5.  7 - 1200 ml stainless steel beakers.
 6.  Several Ibs each of Neoprene stoppers, size 00, 0, and 1, 1 Ib of  siz 5.
 7.  ]4 - aquarium 2-5 gal.
 8.  Magnetic stirrer.
 9.  2 - PVC ball valves, 1/2" pipe size.
10.  Dilutor control panel - see Fig. 14 and equipment list.
11.  Plywood sheeting, exterior grade:  one - 4' x 8'  x 3/4", one -
     4' x 8' x 1/2".
12.  Pine or redwood board, 1" x 8", 20 ft.
13.  Epoxy paint, 1 gal.
14.  Assorted wood screws, nails, etc.
15.  25 ft - 1/4" ID, TEFLONR tubing, to connect the mixing chambers
     to the test chambers (see Fig. 10).
43

-------
                         NORMALLY OPEN
                        ( SOLENOID VALVES
-FLOW CONTROL VALVES

         ^DILUTION INFLOW
                               DILUTION WATER CHAMBERS
                                                       -LIQUID LEVEL SWITCH
                                                         NORMALLY CLOSED
                                                         SOLENOID VALUE
                           TEST CHAMBERS 4-20 LITERS CAPACITY
FIG. 9. VACUUM  SIPHON DILUTOR SYSTEM, GENERAL DIAGRAM.
                                 44

-------
            D-l
                     D-2
                                  D-3
ADJUSTABLE
 STANDPIPE
  DRAIN
DILUTION WATER CHAMBERS

VACUUM LINE

6mm OD CONNECTING TUBE  T  FORM


10 mm OD U SHAPE SYPHON TUBE


6mm 0 D VACUUM LINE TUBE

STAINLESS STEEL HOSE CLAMP


10 mm ID CONNECTING TUBES Y FORM


10 mm 0 D DELIVERY TUBE

120 ml BOTTLE  VACUUM BLOCK


10 mm 0 D. DELIVERY TUBE
                                                    10 mm O.D. DELIVERY TUBE

                                                    10 mm O.D. AUTOMATIC SYPHON TUBE
                                                   EFFLUENT CHAMBERS


                                                   10mm O.D. U SHAPE SYPHON TUBE



                                                   10 mm I.D CONNECTING TUBE Y FORM




                                                   10 mm 0 D. DELIVERY TUBE






                                                   MIXING CHAMBER  I2OO ml CAPACITY





                                                   10 mm O.D. DELIVERY TUBE
                                                    TEST CHAMBERS CAPACITY
                                                    4- 20 LITERS
  FIG. 10. VACUUM SIPHON  DILUTOR SYSTEM, DETAILED DIAGRAM.
                                         45

-------
Vacuum Siphon System Equipment List

 1.  Diluter Glass.
 2.  Stainless steel solenoid valves.
     a.  2 - normally open, two-way, 55 psi, water, 1/4" pipe size,
         9/32" orifice size, ASCO 8262152, for incoming effluent and
         dilution water pipes.
     b.  1 - normally closed, two-way, 15 psi, water, 3/8" pipe size,
         3/8" orifice size, ASCO 8030B65, for dilution water evacuation
         pipe.
3.   Stainless steel tubing, seamless, 316-grade, austenetic.
     a.  60 ft - 3/8" OD, 0.035" wall thickness, for dilution water and
         effluent pipes.
     b.  20 ft - 5/16" OD, 0.035" wall thickness, for standpipes in
         mixing chambers.
     c.  1 ft - 3/4" OD, 0.035" wall thickness, for standpipe in D-l
         chamber.
 4.  Swagelok tube connectors, stainless steel
     a.  4 - male tube connectors, male pipe size 1/4", tube OD 3/8".
     b.  2 - male tube connectors, male pipe size 3/8", tube OD 3/8".
     c.  2 - male adapter, tube to pipe, male pipe size 1/2", tube OD 3/8",
     d.  2 - male tube connectors, male pipe size 1/2", tube OD 3/8".
 5.  7 - 1,200 ml stainless steel beakers.
 6.  Several Ibs each of NEOPRENER stoppers, size 00, 0 and 1; 1 Ib of
     size 5.
 7.  14 - aquariums, 2-5 gal.
 8.  Magnetic stirrer.
 9.  2 - PVC Ball valves, 1/2" pipe size.
10.  Dilutor control panel equipment - see Fig. 14 and equipment list.
11.  7 - 120 ml NALGENER bottles.
12.  3 ft, l-in-2 aluminum bar, for siphon support brackets.
13.  Stainless steel set screws, box of 50, for securing SS tubing in
     siphon support brackets.
14.  Stainless steel hose clamps, box of 10, size #4 or 5,(need 3 boxes).
15.  6 - NALGENER T's, 5/16" OD.
16.  12 - TYGONR Y connectors, 3/8" I.D.
17.  TYGONR tubing, 3/8" OD, 10 ft.
18.  Plywood sheeting, exterior grade:  one - 4' x 8x x 3/4",
     one - 4' x 8' x 1/2".
19.  Pine or redwood board, 1" x 8", 20 ft.
20.  Epoxy paint, 1 gal.
21.  Assorted wood screws, nails, etc.
22.  25 ft - 5/16" ID, TEFLONR tubing, to connect the mixing chambers to
     the test chambers.
                                     46

-------
 231 m m
A
iK N
D-l

D

B
(.
— 1
-2




D-3

B
D

4





D-5


B
D-6




D-7
1

L
D-8

                              609 m m
                20mm  HOLE  CENTERED AT INDICATED DISTANCE
                                                             x 20 mm HOLE FOR VACUUM
                                                          .. Ll SIPHON SYSTEM ONLY.
                                                          -  - (See Fig. 6)
           oo       ooooooo
         33mm 70mm    165mm  248mm 297mm 350mm 411 mm 477mm  558mm

DRAIN HOLES IN BOTTOM PLATE 1C) SHOWN FOR SOLENOID VALVE DILUTOR SYSTEM. FOR VACUUM
SIPHON DILUTOR SYSTEM, SINGLE DRAIN HOLE IS REQUIRED ONLY FOR CHAMBER D-1.
                  INDIVIDUAL PART SIZE AND NUMBER OF
                  PIECES  USING  6 mm (I/4")PLATE GLASS
                         A  225 mm X  95 mm - 3
                         B  200 mm X  95mm - 6
                         C  609 mm X  95mm - I  (Bottom Plate)
                         D  609 mm X 23! mm - 2 (Side Panels)

          INSIDE CELL MEASUREMENTS AND APPROXIMATE VOLUME
                    D-l 95 mm  X 225 mm X 95 mm -2030ml
                    D - 2 IIS mm  X 200 mm X 95 mm -2185ml
                    D-3 40 mm  X 200mm X 95 mm - 760ml
                    D-4 45mm  X 200 mm X 95mm- 855ml
                    D-5  50 mm  X 200 mm X 95mm - 950 ml
                    D-6  60 mm  X 200 mmX 95mm - 114 Oml
                    D-7  60mm  X 200 mmX 95mm-ll40ml
                    D-8  90mm  X 200mm X 95mm-l7IOml
                  NOTE: 1/8" - 316 GRADE AUSTENISTIC STAINLESS STEEL
                       MAY BE SUBSTITUTED FOR GLASS IN PART C

           FIG. II. DILUTION  WATER CHAMBERS
                                        47

-------
180 mm
] E
E-l
. 	 |_i 	
j E
E-2
	 U 	
3 E
E-3
	 M —
i E
A
E-4
-— M—
i e
E-5

"
E-6
I I ... „
t 	 -y 	 j-i 	 H 	 N 	 h 	 	
                                                                      20 mm HOLE
                                                                      FOR VACUUM
                                                                      SIPHON SYSTEM ONLY.
                                                                      (See Fig. 6)
                                   367 mm
                    20 mm HOLE CENTERED AT INDICATED DISTANCE
                     o
o      o    o   o     o
                   61mm
                                  154 mm
                                           213 mm 254mm  290mm 336mm
   DRAIN HOLES IN BOTTOM PLATE (C) SHOWN FOR SOLENOID VALVE DILUTOR SYSTEM. FOR VACUUM
   SIPHON DILUTOR SYSTEM, SINGLE DRAIN HOLE IS REQUIRED ONLY FOR CHAMBER D-1.
                       INDIVIDUAL PART SIZE AND  NUMBER OF
                       PIECES  USING 6mm (l/4")PLATE GLASS

                             A  I74 mm  X 40 mm - Z

                             B  I55 mm  X 40 mm - 5
                             C  367mm  X 40 mm -  I (Bottom Plate)

                             0  367mm  X  180mm-  2 (Side Panels)

              INSIDE  CELL MEASUREMENTS AND APPROXIMATE  VOLUME


                       E-l  NOmm X 155mm X40 mm - 682 ml
                       E- 2  65 mm X 155 mm X 40 mm - 403 ml

                       E-3  40 mm X l55mmX40mm-248ml
                       E;4  30mm X !55mmX  40mm- 186ml
                       E-5  30 mm X l55mmX40mm- 186ml
                       E-6  50mm XI55mmX40mm-3l8ml
                  NOTE: 1/8" - 316 GRADE AUSTENISTIC STAINLESS STEEL
                       MAY BE SUBSTITUTED FOR GLASS IN PART C.
                   FIG.  12.  EFFLUENT  CHAMBERS
                                         48

-------
      M 2
186mm
                 SIDE  VIEW
                                                 M-l
                        I 90 mm
                                            -M-3
        180 m m
                      PART M-l
                   END VIEW
                                        20 mm  DIAMETER HOLE
                      105 mm
            INDIVIDUAL PART SIZE AND  NUMBER OF
            PIECES USING  6mm (I/4")PLATE GLASS

                  M- I  180mm X 105 mm-l
                  M -Z 180mm X 105 mm-l
                  M-3 190mm X 105 mm-1 (BOTTOM PLATE)
                  M- 4 190mm X !86mm-2 (SIDE PANELS)

            APPROXIMATE   CAPACITY  3365 ml )

             FIG. 13. MIXING CHAMBER
                           49

-------
                                                              TDR-I
ALL SOLENOIDS
                           FIG. 14. DILUTOR CONTROL PANEL.

-------
Dilutor Control Panel Equipment List*
Designation


   Al

   CTR-1

   ET

   F,
    2

   L.S.
   SJ,
   SJ,,
CRT Description

Encapsulated amplifier

Cycle counter

Elapsed time indicator

Input power fuse

Recepticle

Aux A.C. output jack

Main input power cord

Fill indicator light

Emptying indicator light

Level sensor (Dual Sensing Probe)

Plug

On-off main power switch (spst)

On-off aux power  switch (spst)

Solenoid
      Manufacturer

Cutler Hammer 1353H98C

Rodington #P2-1006

Courac #636W-AA H&T

Little fuse 342038

Amphenol 91PC4F

Stand. 3-prong AC Recpt.

Stand. 3-prong AC maleplug

Dialco 95-0408-09-241

Dialco 95-0408-09-241

Cutler Hammer 13653H2

Amphenol 91MC4M

Cutler Hammer 7580 K7

Cutler Hammer 7580 K7

(See Solenoid and Vacuum
 System equipment lists)
SJ. - SJ,,
4 16
TDR-1
TDK- 2
Additional Solenoids for
Solenoid Valve System
Time delay relay
Aux time delay relay
ii it M
Dayton 5x829
Dayton 5x829
*Consult local electric supply house.
                                     51

-------
                                   TECHNICAL REPORT DATA
                            (Please read Instructions on the reverse before completing)
1. REPORT NO.
 EPA-600/4-78-012
                              2.
                                                            3. RECIPIENT'S ACCESSIOI*NO.
4. TITLE AND SUBTITLE
   METHODS  FOR MEASURING  THE ACUTE TOXICITY  OF
   EFFLUENTS TO AQUATIC ORGANISMS
                                     5. REPORT DATE
                                       January 1978  issuing date
                                     6. PERFORMING ORGANIZATION CODE
7. AUTHOR(S)
   William Peltier
                                                            8. PERFORMING ORGANIZATION REPORT NO.
9. PERFORMING ORGANIZATION NAME AND ADDRESS
   Bioassay Subcommittee
   EPA Biological Advisory Committee
   Environmental Research Laboratory
   Athens,  Georgia 30601
                                     10. PROGRAM ELEMENT NO.

                                             1BD612
                                     11. CONTRACT/GRANT NO.
12. SPONSORING AGENCY NAME AND ADDRESS
   Environmental Monitoring & Support Lab.- Gin., OH
   Office of Research  and Development
   U.S.  Environmental  Protection Agency
   Cincinnati, Ohio 45268
                                     13. TYPE OF REPORT AND PERIOD COVERED
                                       In-House
                                     14. SPONSORING AGENCY CODE

                                            EPA/600/06
15. SUPPLEMENTARY NOTES
        view i A*n T raw i co
         Supplement to  "Biological Field  and Laboratory Methods for Measuring the
    Quality of Surface  Waters & Effluent"
16. ABSTRACT
               This report describes methods for the measurement of the acute
    toxicity of effluents  to macroinvertebrates and fish.   The methods include a
    preliminary short-term (8-24 hr.)» range-finding test  and a long-term  (96 hr.),
    or alternate static  definitive test  for use in determining the LC50 or  EC50 of
    the waste.  The report includes guidelines on effluent sampling and holding,
    facilties and equipment, dilution water, test species  selection and handling,
    and data interpretation.
17.
                                KEY WORDS AND DOCUMENT ANALYSIS
                  DESCRIPTORS
                                              b. IDENTIFIERS/OPEN ENDED TERMS
                                                   c. COSATI Field/Group
    Effluents
    Bioassay
    Toxicity
    Industrial Wastes
    Sewage
    Water Pollution
    Fishes
Invertebrates
Fresh Water  Biology
Marine Biology
                                                      6C
18. DISTRIBUTION STATEMENT
       RELEASE  TO PUBLIC
                        19. SECURITY CLASS (ThisReport)
                            UNCLASSIFIED
21. NO. OF PAGES
     62
                                               2O. SECURITY CLASS (Thispage)
                                                  UNCLASSIFIED
                                                                          22. PRICE
EPA Form 2220-1 (9-73)
                                             52
                                                          S. GOVERNMENT PRINTING OFFICE. 1378-757-1W6615 Region No.  5"

-------