<   *       EPA 600/4-81-056
                 United States
                 Environmental Protection
                 Agency
vvEPA         Research  and
                 Development
                     Jgtal  Organic Halide
                     Method  450.1 - Interim
                 Prepared for
                  Joseph A. Cotruvo
                  Director
                  Criteria and Standards Division
                  Office of Drinking Water
                 Prepared by
                  Stephen Billets, Ph.D.
                  James J. Uchtenberg
                  Physical and Chemical Methods Branch
                  Environmental Monitoring and Support Laboratory
                  Cincinnati, Ohio 45268

-------
              TOTAL  ORGANIC HALIDE
                  Method 450.1

                    Interim
     U  S. Environmental Protection Agency
      Office  of  Research and Development
Environmental  Monitoring and Support Laboratory
      Physical  and  Chemical  Methods Branch
            Cincinnati, Ohio  45268
                 November 1980


             U.S. Environ-,-,-•  ; ~    :^:on Agency
             Region  V, L ;
             230 Soul'i  'j      ••,'•••  -jr
                             ' J • V-  ' .  •>*• -f tt
             Chicago, liii.^-j  COo-04

-------
                             TOTAL ORGANIC HALIDE
                                 Method 450.1

1.  Scope and Application
    1.1  This method is to be used for the determination of Total Organic
         Hal ides as Cl" by carbon adsorption, and requires that all
         samples be run in duplicate.  Under conditions of duplicate
         analysis, the reliable limit of sensitivity is 5 ug/L.  Organic
         halides as used in this method are defined as all organic species
         containing chlorine, bromine and iodine that are adsorbed by
         granular activated carbon under the conditions of the method.
         Fluorine containing species are not determined by this method.
    1.2  This is a microcoulometric-titration detection method applicable to
         the determination of the compound class listed above in drinking
         and ground waters, as provided under 40 CFR 265.92.
    1.3  Any modification of this method, beyond those expressly permitted,
         shall be considered as major modifications subject to application
         and approval of alternate test procedures under 40 CFR 260.21.
    1.4  This method is restricted to use by, or under the supervision of,
         analysts experienced in the operation of a pyrolysis/microcolumeter
         and in the interpretation of the results.
2.  Summary of Method
    2.1  A sample of water that has been protected against the loss of
         volatiles by the elimination of headspace in the sampling
         container, and is free of undissolved solids, is passed through a
         column containing 40 mg of activated carbon.  The column is washed

-------
         to remove any trapped inorganic halides, and is then pyrolyzed to
         convert the adsorbed organohalides to a titratable species that can
         be measured by a microcoulometric detector.
3.  Interferences
    3.1  Method interferences may be caused by contaminants, reagents,
         glassware, and other sample processing hardware.  All of these
         materials must be routinely demonstrated to be free from
         interferences under the conditions of the analysis by running
         method blanks.
         3.1.1   Glassware must be scrupulously cleaned.  Clean all glassware
                as soon as possible after use by treating with chromate
                cleaning solution.   This should be followed by detergent
                washing in hot water.   Rinse with tap water and distilled
                water, drain dry, and  heat in a muffle furnace at 400°C
                for 15 to 30 minutes.   Volumetric ware should not be heated
                in a muffle furnace.  Glassware should be sealed and stored
                in a clean environment after drying and cooling, to prevent
                any accumulation of dust or other contaminants.
         3.1.2   The use of high purity reagents and gases help to minimize
                interference problems.
    3.2  Purity of the activated carbon must be verified before use.  Only
         carbon samples which register less than 1000 ng/40 mg should be
         used.   The stock of activated carbon should be stored in its
         granular form in a glass container with a Teflon seal.  Exposure to
         the  air must be minimized,  especially during and after milling and
         sieving the activated carbon.  No more than a two-week supply

-------
         should be prepared in advance.  Protect carbon at all times from
         all sources of halogenated organic vapors.  Store prepared carbon
         and packed columns in glass containers with Teflon seals.
    3.3  This method is applicable to samples whose inorganic-halide
         concentration does not exceed the organic-halide concentration by
         more than 20,000 times.
4.  Safety
    The toxicity or carcinogenicity of each reagent in this method has not
    been precisely defined; however, each chemical compound should be
    treated as a potential health hazard.  From this viewpoint, exposure to
    these chemicals must be reduced to the lowest possible level by whatever
    means available.  The laboratory is responsible for maintaining a
    current-awareness file of OSHA regulations regarding the safe handling
    of the chemicals specified in this method.  A reference file of
    material-handling data sheets should also be made available to all
    personnel involved in the chemical analysis.
5.  Apparatus and Materials  (All specifications are suggested.  Catalog
    numbers are included for illustration only).
    5.1  Sampling equipment, for discrete or composite sampling
         5.1.1  Grab-sample bottle - Amber glass, 250-ml, fitted with
                Teflon-lined caps.  Foil may be substituted for Teflon if
                the sample is not corrosive.  If amber bottles are not
                available, protect samples from light.  The container must
                be washed and muffled at 400°C before use, to minimize
                contamination.

-------
5.2  Adsorption System
     5.2.1  Dohrmann Adsorption Module  (AD-2), or equivalent,
            pressurized, sample and nitrate-wash reservoirs.
     5.2.2  Adsorption columns - pyrex, 5 cm long X 6-mm OD X 2-mm  ID.
     5.2.3  Granular Activated Carbon (GAC) - Filtrasorb-400,
            Calgon-APC, or equivalent,  ground or milled, and screened to
            a 100/200 mesh range.  Upon combustion of 40 mg of GAC, the
            apparent-halide background  should be 1000-mg Cl"
            equivalent or less.
     5.2.4  Cerafelt (available from Johns-Manville), or equivalent -
            Form this material into plugs using a 2-mrn ID
            stainless-steel borer with  ejection rod (available from
            Dohrmann) to hold 40 mg of  GAC in the adsorption columns.
            CAUTION:  Do not touch this material with your fingers.
     5.2.5  Column holders (available from Dohrman).
     5.2.6  Volumetric flasks - 100-mL, 50-mL.
            A general schematic of the  adsorption system is shown in
            Figure 1.
5.3  Dohrmann microcoulometric-titration system (MCTS-20 or DX-20), or
     equivalent, containing the following components:
     5.3.1  Boat sampler.
     5.3.2  Pyrolysis furnace.
     5.3.3  Microcoulometer with integrator.
     5.3.4  Titration cell.
            A general description of the analytical system is shown in
            Figure 2.
5.4  Strip-Chart Recorder.

-------
6.  Reagents
    6.1  Sodium sulfite - 0.1 M, ACS reagent grade (12.6 g/L).
    6.2  Nitric acid - concentrated.
    6.3  Nitrate-Wash Solution (5000 mg NOg/L) - Prepare a nitrate-wash
         solution by transferring approximately 8.2 gm of potassium nitrate
         into a 1-litre volumetric flask and diluting to volume with reagent
         water.
    6.4  Carbon dioxide - gas, 99.9% purity.
    6.5  Oxygen - 99.9% purity.
    6.6  Nitrogen - prepurified.
    6.7  70% Acetic acid in water - Dilute 7 volumes of acetic acid with 3
         volumes of water.
    6.8  Trichlorophenol solution, stock (1 yL = 10 yg Cl") - Prepare a
         stock solution by weighing accurately 1.856 gm of trichlorophenol
         into a 100-mL volumetric flask.  Dilute to volume with methanol.
    6.9  Trichlorophenol solution, calibration (1 yL = 500 ng Cl") -
         Dilute 5 ml of the trichlorophenol stock solution to 100 ml with
         methanol.
    6.10 Trichlorophenol standard, instrument-calibration - First, nitrate
         wash a single column packed with 40 mg of activated carbon as
         instructed for sample analysis, and then inject the column with
         10 yL of the calibration solution.
    6.11 Trichlorophenol standard, adsorption-efficiency (100 yg C1"/L) -
         Prepare a adsorption-efficiency standard by injecting 10 yL of
         stock solution into 1 liter of reagent water.
    6.12 Reagent water - Reagent water is defined as a water in which an

-------
         interferent is not observed at the method detection limit of each
         parameter of interest.
    6.13 Blank standard - The reagent water used to prepare the calibration
         standard should be used as the blank standard.
7.  Calibration
    7.1  Check the adsorption efficiency of each newly-prepared batch of
         carbon by analyzing 100 ml of the adsorption-efficiency standard,
         in duplicate, along with duplicates of the blank standard.  The net
         recovery should be within 5% of the standard value.
    7.2  Nitrate-wash blanks (Method Blanks) - Establish the repeatability
         of the method background each day by first analyzing several
         nitrate-wash blanks.  Monitor this background by spacing nitrate-
         wash blanks between each group of eight pyrolysis determinations.
         7.2.1  The nitrate-wash blank values are obtained on single columns
                packed with 40 mg of activated carbon.  Wash with the
                nitrate solution as instructed for sample analysis, and then
                pyrolyze the carbon.
    7.3  Pyrolyze duplicate instrument-calibration standards and the blank
         standard each day before beginning sample analysis.  The net
         response to the calibration-standard should be within 3% of the
         calibration-standard value.  Repeat analysis of the
         instrument-calibration standard after each group of eight pyrolysis
         determinations, and before resuming sample analysis after cleaning
         or reconditioning the titration cell or pyrolysis system.
8.  Sample Preparation
    8.1  Special care should be taken in the handling of the sample to

-------
         minimize the loss of volatile organohalides.  The adsorption
         procedure should be performed simultaneously on duplicates.
    8.2  Reduce residual chlorine by the addition of sulfite (1 ml of 0.1 M
         per liter of sample).  Addition of sulfite should be done at the
         time of sampling if the analysis is meant to determine the TOX
         concentration at the time of sampling.  It should be recognized
         that TOX may increase on storage of the sample.  Samples should be
         stored at 4°C without headspace.
    8.3  Adjust pH of the sample to approximately 2 with concentrated HNOg
         just prior to adding the sample to the reservoir.
9.  Adsorption Procedure
    9.1  Connect two columns in series, each containing 40 mg of
         100/200-mesh activated carbon.
    9.2  Fill the sample reservoir, and pass a metered amount of sample
         through the activated-carbon columns at a rate of approximately
         3 mL/min.  NOTE:  100 ml of sample is the preferred volume for
         concentrations of TOX between 5 and 500 ug/L; .50 ml for 501 to 1000
         yg/L, and 25 ml for 1001 to 2000 ug/L.
    9.3  Wash the columns-in-series with 2 mL of the 5000-mg/L nitrate
         solution at a rate of approximately 2 mL/min to displace inorganic
         chloride ions.
10.  Pyrolysis Procedure
    10.1 The contents of each column is pyrolyzed separately.  After rinsing
         with the nitrate solution, the columns should be protected from the
         atmosphere and other sources of contamination until ready for
         further analysis.

-------
    10.2 Pyrolysis of the sample is accomplished in two stages.  The
         volatile components are pyrolyzed in a COp-rich atmosphere at a
         low temperature to assure the conversion of brominated
         trihalomethanes to a titratable species.  The less volatile
         components are then pyrolyzed at a high temperature in an CL-rich
         atmosphere.
         NOTE:   The quartz sampling boat should have been previously muffled
         at 800 C for at least 2 to 4 minutes as in a previous analysis,
         and should be cleaned of any residue by vacuuming.
    10.3 Transfer the contents of each column to the quartz boat for
         individual analysis.
    10.4 If the Dohrmann MC-1 is used for pyrolysis, manual instructions are
         followed for gas flow regulation.  If the MCT-20 is used, the
         information  on the diagram in Figure 3 is used for gas flow
         regulation.
    10.5 Position the sample for 2 minutes in the 200°C zone of the
         pyrolysis tube.  For the MCTS-20, the boat is positioned just
         outside the  furnace entrance.
    10.6 After  2 minutes, advance the boat into the 800°C zone (center) of
         the pyrolysis furnace.  This second and final stage of pyrolysis
         may require  from 6 to 10 minutes to complete.
11.  Detection
    The effluent gases are directly analyzed in the microcoulometric-titra-
    tion cell.   Carefully follow manual instructions for optimizing cell
    performance.

-------
12.  Breakthrough
    Because the background bias can be of such an unpredictable nature, it
    can be especially difficult to recognize the extent of breakthrough of
    organohalides from one column to another.  All second-column
    measurements for a properly operating system should not exceed
    10-percent of the two-column total measurement.  If the 10-percent
    figure is exceeded, one of three events can have happened.  Either the
    first column was overloaded and a legitimate measure of breakthrough was
    obtained - in which case taking a smaller sample may be necessary; or
    channeling or some other failure occurred - in which case the sample may
    need to be rerun; or a high, random,  bias occurred and the result should
    be rejected and the sample rerun.  Because knowing which event has
    occurred may not be possible, a sample analysis should be repeated often
    enough to gain confidence in results.  As a general rule, any analyses
    that is rejected should be repeated whenever sample is available.  In
    the event that the second-column measurement is equal to or less than
    the nitrate-wash blank value, the second-column value should be
    disregarded.
13.  Quality Control
    13.1 Before performing any analyses,  the analyst must demonstrate the
         ability to generate acceptable accuracy and precision with this
         procedure by the analysis of appropriate quality-control check
         samples.
    13.2 The laboratory must develop and maintain a statement of method
         accuracy for their laboratory.  The laboratory should update the
         accuracy statement regularly as  new recovery measurements are made.

-------
    13.3 It is recommended that the laboratory adopt additional
         quality-assurance practices for use with this method.  The specific
         practices that would be most productive will depend upon the needs
         of the laboratory and the nature of the samples.  Field duplicates
         may be analyzed to monitor the precision of the sampling
         technique.  Whenever possible, the laboratory should perform
         analysis of standard reference materials and participate in
         relevant performance-evaluation studies.
14. Calculations
    OX as Cl~ is calculated using the following formula:
                    (cr c3) + (c2 - c3 ) u  ug/L Total Organ1c Ha-|ide
    where:
    C, = ug Cl~ on the first column in series
    C~ - ug Cl" on the second column in series
    Co = predetermined, daily, average, method-blank value
           (nitrate-wash blank for a 40-mg carbon column)
    V = the sample volume in L
15. Accuracy and Precision
    These procedures have been applied to a large number of drinking-water
    samples.  The results of these analysis are summarized in Tables I and
    II.
16. Reference
    Dressman, R., Najar, G., Redzikowski, R., paper presented at the
    Proceedings of the American Water Works Association Water Quality
    Technology Conference, Philadelphia,  Dec. 1979.

-------
                    TABLE I



PRECISION AND ACCURACY DATA FOR MODEL COMPOUNDS
Model Dose Dose Average Standard
Compound ug/L as yg/L Cl % Recovery Deviation
CHC13 98 88 89 14
CHBrCl2 160 106 98 9
CHBr2C1 155 79 86 11
CHBr3 160 67 111 8
Pentachlorophenol 120 80 93 9
TABLE II
PRECISION DATA ON TAP WATER ANALYSIS
Avg. halide Standard
Sample ug Cl/L Deviation
A 71 4.3
B 94 7.0
C 191 6.1
No. Of
Replicates
10
11
13
11
7
No. of
Replicates
8
6
4

-------
I\l2
  SAMPLE
RESERVOIR
  (1 of 4)
NITRATE WASH
RESERVOIR
 GAG COLUMN 1
 GAC COLUMN 2
           Figure 1. Adsorption Schematic

-------
              SPARGING
               DEVICE
TITRATION
pel I



PYROLYSIS
FURNACE



BOAT
INLET
MICROCOULOMETER
 WITH INTEGRATOR
STRIP CHART
 RECORDER
                                         ADSORPTION
                                           MODULE
   Figure 2. CAOX Analysis System Schematic

-------
                                     B
                                             PYROLYSIS FURNACE



SINGLE BOAT OUTLET

CONNECT TO
BOAT-INLET
PORTS A

II 00
i |l| I"' *V^^y^~|
4- II 1
543 21
IIOiVENT CAPPEDI
f «












1
1





H













.sx r n
N.Al .
' * * i 	 \ <
i ' «
Vr\ LI
0 c— ' | CO. 50 ml/min O, 100 ml/min





| CARRIER OUT REACTANT OUT
! o o
I
I
I
L_Un
I ^

i
r



COi 100 ml/min
AUXILIARY
OUT

1






1

Figure 3. Rear view plumbing schematic for MCTS-20 system.
         Valve A is set for first-stage combustion. O2 venting
         (push/pull valve out). Port B enters inner combustion
         tube; Port C enters outer combustion tube.

-------