United States                             __. „ _„ ,. „„ „„„
                  Environmental Protection                       EPA# 600/4-81-059
                  Agency
v>EPA         Research and
                  Development
                  The Determination of Halogenated Chemicals
                               in Water
                        by the Purge and Trap Method
                             Method 502.1
                  Prepared for
                           Joseph A. Cotruvo
                           Director
                           Criteria and Standards Division
                           Office of Drinking Water
                  Prepared by
                          Thomas A. Beliar
                          James J. Lichtenberg
                          Physical and Chemical Methods Branch
                          Environmental Monitoring and Support Laboratory
                          Cincinnati, Ohio 45268
                                            us,

-------
                    Determination of Halogenated Chemicals
                    in Water by the Purge and Trap Method'
                                 Method 502.1
1.0 SCOPE AND APPLICATION
    1.1  This method is applicable to the determination of carbon
         tetrachloride, 1,2-dichloroethane, tetrachloroethylene, 1,1,1,-tri-
         chloroethane, trichloroethylene and vinyl chloride contained in
         finished drinking water, raw source water, and such water in any
         state of treatment.  These compounds and an additional 44 halo-
         genated compounds which can be determined by this method are listed
         in Table I.
    1.2  Single laboratory accuracy and precision data show that this proce-
         dure is useful for the detection and measurement of multicomponent
         mixtures spiked.into carbon filtered finished water and raw source
         water at concentrations between 0.20 and 0.40 ug/L with method
         detection limits generally less than 0.01 ug/L..  The method as
         described is capable of accurately measuring those compounds men-
         tioned in Table I over a concentration range of 0.10 to 5.0 ug/L.
         Additionally, it is possible to measure individual organohalides up
         to 1500 ug/L.  However the ability to measure complex mixtures
         containing co-eluting or partially resolved
1  Organic Analyses Section,  Environmental  Monitoring and Support
    Laboratory,  Cincinnati,  Ohio 45268,  April  1981,

-------
         organohaTides with concentration differences larger than  a factor
         of 10 is hampered.  This problem commonly occurs when finished
         drinking waters are analyzed because of the relatively high
         trihalomethane content.  When such samples are analyzed chloroform
         will  affect the method detection limit for 1,2-dichloroethane.
    1.3  This  method is recommended for use only by analysts experienced in
         the measurement of purgeable organics at the low ug/L level or by
         experienced technicians under the close supervision of a  such
         qualified analyst.
2.  Summary
    2.1  An extraction/concentration technique is incorporated within the
         method which enhances the quantities of organohalides injected into
         the gas chromatograph by a factor of 1000 over direct injection gas
         chromatography.
    2.2  Organohalides are extracted by an inert gas which is bubbled
         through the aqueous sample.  The organohalides, noted in Table I
         along with other organic constituents which exhibit low water
         solubility and boil less than 200°C, are efficiently transferred
         from  the aqueous phase to the gaseous phase.   These compounds are
         swept from the purging device and are trapped in a short column
         containing a carefully selected sorbant combination.   After a
         predetermined period of time, the trapped components are thermally
         desorbed and backflushed onto the head of a gas chromatographic
         column and separated under programmed conditions.
    2.3  Measurement is accomplished with a halogen specific detector which
         eliminates interference problems commonly encountered with
         universal  or semispecific detectors.   Under special conditions

-------
         (4.2.2c), a mass spectrometer may be used in place of a halogen
         specific detector for quantitative measurement.
    2.4  Confirmatory analyses are performed using dissimilar columns.  If
         sufficient material is present, confirmatory analyses are performed
         by gas chromatography-mass spectrometry (GC/MS).
    2.5  Aqueous standards and unknowns are extracted and analyzed under
         identical conditions to compensate for extraction losses (See Table
         1 for the purging efficiency of individual compounds).
    2.6  To minimize personal exposure to known human carcinogens, an alter-
         native non-aqueous calibration procedure is provided for vinyl
         chloride.  The gas chromatograph is calibrated by injecting dilute
         commercially available gaseous standards of vinyl chloride into the
         purging device as aqueous standards of other compounds are purged.
    2.7  The total analysis time is approximately 1 hour per sample.
3.  Interferences
    3.1  During the development and testing of this method, certain analyti-
         cal parameters and equipment design were found to affect the valid-
         ity of the analytical results.  Proper use of the method requires
         that such parameters or design be used as specified.  These items
         are identified in the text by the word "must."  Anyone wishing to
         deviate from the method on these operations must demonstrate that
         such a deviation does not affect the validity of the data.  Alter-
         native test procedure approval must be obtained through the
         Environmental Monitoring and Support Laboratory - Cincinnati
         Equivalency Program.  An experienced analyst may make modifications
         to parameters or equipment not identified by the word "must" as
         long as the data obtained are equivalent to or better than that

-------
     obtained with the method as presented in Section 9. Formal approval
     is not required, but documented comparable data must be on file as
     part of the overall quality assurance program.
3.2  Impurities contained in the purge- gas and organic compounds out
     gasing from the plumbing ahead of the trap usually account for the
     majority of contamination problems.  The presence of such inter-
     ferences is easily monitored using the quality control program
     described herein.  Field reagent blanks (FRB) are normally run
     between each set (6.4.9.5) of samples. When a positive organohalide
     response is noted in the FRB, the analyst should analyze a
     laboratory reagent blank (LRB) inorder to identify the source of
     contamination.  LRB are run by charging the purging device with
     reagent water and analyzing it in the normal manner. Whenever
     organohalides are noted in the LRB, the analyst should change the
     purge gas source and regenerate the molecular sieve purge gas
     filter.  Subtracting blank values from sample results is not
     recommended.  The use of non-TFE plastic tubing, non-TFE thread
     sealants, or flow controllers with rubber components in the purging
     device should be avoided since such materials out-gas organic
     compounds which will be concentrated in the trap during the purge
     operation.  Such out-gassing problems are common whenever new
     equipment is put into service.  With use, minor out-gassing
     problems generally cure themselves.
3.3  Several instances of accidental sample contamination have been
     noted and attributed to diffusion of volatile organics through the
     septum seal and into the sample during shipment and storage.  The
     FRB is used as a monitor for this problem.  If the FRB is contain-

-------
     inated according to 9.11 the entire sample set must be discarded
     and resampled.
3.4  For compounds that are not efficiently purged, such as tetrachloro-
     ethylene, small variations in sample volume, purge time, purge flow
     rate, purging device geometry, or purge temperature can affect the
     analytical result.  Therefore, with the exception of vinyl
     chloride, samples and standards must be analyzed under identical
     conditions.
3.5  In cases of compounds with low boiling points, such as vinyl
     chloride, small variations in purging conditions and trap configur-
     ation can cause trap saturation and sample venting.  For this
     reason, sample volume, purging conditions and trap parameters must
     be duplicated within the constraints of this method.  Variations
     are permitted only through the EMSL-Cincinnati Equivalency Program.
3.6  Cross-contamination can occur whenever high level and low level
     samples are sequentially analyzed.  To reduce the likelihood of
     this, the purging device and sample syringe must be rinsed at least
     twice between samples with reagent water.  Whenever an unusually
     concentrated sample is encountered, it is necessary that it be fol-
     lowed by FRB or LRB analysis to check for sample cross contamina-
     tion.  Note:  Cross-contamination problems vary between instruments.
     It is up to the analyst to determine when this practice is
     required.  For samples containing large amounts of water soluble
     materials, suspended solids, high boiling compounds or high organo-
     halide levels it may be necessary to wash out the purging device
     with a soap solution, rinse with distilled water, and then dry in a
     105°C oven between analyses.

-------
    3.7  Qualitative misidentifications can occur in gas chromatographic
         analysis.   Whenever samples whose qualitative nature is unknown are
         analyzed,  the following precautionary measures should be
         incorporated into the analysis.
         3.7.1  Perform duplicate analyses using the two recommended columns
               (4.2.1.1 and 4.2.1.2) which provide different retention order
               and  retention times for many organohalides.
         3.7.2  Whenever possible use GC/MS techniques which provide
               unequivocal qualitative identifications.
    3.3  To analyze for methylene chloride at concentrations below 1 ug/L
         the laboratory will need to take the following special precautions.
         The analytical and sample storage area should be isolated from all
         atmospheric sources of methylene chloride, otherwise random back-
         ground levels will result.  Methylene chloride will permeate
         through Teflon tubing.  For this reason, all GC carrier gas lines
         and purge  gas plumbing should be constructed from stainless steel
         or copper  tubing.  Laboratory clothing exposed to methylene
         chloride fumes during common liquid/liquid extraction procedures
         can contribute to sample contamination during steps described in
         Sections 5, 6, and 8.
4.  Apparatus
    4.1  The purge and trap equipment consists of three separate pieces of
         apparatus:  the purging device, the trap, and the desorber.
         Construction details for a purging device and an easily automated
         trap-desorber hybrid that was used to generate the single
         laboratory accuracy and precision data listed in Section 11 is
         shown in Figures 1 through 4 and described  in 4.1.1 through 4.1.3.

-------
4.1.1  Purging Device -- Construction details are given in Figure 1
       for an acceptable all glass 5 ml purging device.  The glass
       frit installed at the base of the sample chamber allows
       finely divided gas bubbles to pass through the sample while
       the sample is restrained above the frit.  Gaseous volumes
       above the sample are kept to a minimum to eliminate dead
       volume effects, yet allowing sufficient space for most foams
       to disperse.  The inlet and exit ports are constructed from
       heavy walled 1/4 inch glass tubing so that leak-free
       removable connections can be made using "finger-tight"
       compression fittings containing Teflon ferrules.  The
       removable foam trap is used to control samples that foam.
       The purging device must meet the following specifications:
       a. The purging device must be designed to accept 5 ml
          samples with a water column at least 3 cm deep.
       b. The gaseous head space between the water column and the
          trap must have a total volume of  less than 15 ml.
       c. The purge gas must pass through the water column as
          finely divided bubbles with a diameter of less than 3 mm
          at the origin.
       d. The purge gas must be introduced  no more than 5 mm from
          the base of the water column.
4.1.2  Trapping Device — The trap (Figure  2) is a short gas
       chromatographic column which, at 22°C, retards the flow of
       the compounds of interest while venting the purge gas.  A
       trap is constructed with a low thermal mass so that it can
       be rapidly heated for efficient desorption, then rapidly

-------
cooled to room temperature for recycling.  The 3% OV-1 area
of the trap is used primarily as a thermal spacer to ensure
that the Tenax is contained in a heated area of the trap.
In addition, it isolates the Tenax area of the trap from
high boiling aerosols which may adversely affect the
performance of the Tenax.

The Tenax portion of the trap, under the given operating
conditions, quantitatively retains those compounds listed in
Table 1 from dichloromethane through p-dichlorobenzene
(compounds boiling above 35°C).  The silica gel quantita-
tively retains chloromethane, bromomethane, vinyl chloride
and chloroethane and the charcoal quantitatively retains
dichlorodifloromethane.

Pack the trap according to Figure 2.  In order to function
properly the trap must be constructed so that it meets or
exceeds the sorbant dimentions shown in Figure 2.
Additional requirements and acceptable modifications are as
follows:
a. Place the glass wool plug in the inlet end of the trap,
   follow with the OV-1, Tenax, silica gel, charcoal, and
   finally, the second glass wool plug.  Reversing the
   packing order (placing the charcoal in the trap first)
   will cause the silica gel and Tenax layers to become
   contaminated with charcoal dust causing poor desorption
   efficiencies.
b. The trap must be installed so that the effluent from the
   purging device enters the Tenax end of the trap.

-------
       c. If it is not necessary to analyze for dichlorodifloro-
          methane, the trap packing can be modified by eliminating
          the charcoal and adding additional Tenax and silica gel.
          In this situation, the packing order and dimensions of
          the sorbents in the trap are as follows:  5 mm glass
          wool, 1 cm 3% OV-1, 15 cm Tenax, 8 cm silica gel and 5 mm
          glass wool.
       d. If only compounds boiling above 35°C are to be
          analyzed, the silica gel and charcoal area of the trap
          can be replaced with Tenax providing a Tenax column of 23
          cm.
4.1.3  Desorber assembly — Details for the desorber are shown in
       Figures 3 and 4.  With the 6-port valve in the Purge-Sorb
       position (See Figure 3), the effluent from the purging
       device passes through the trap where the flow rate of the
       organics is retarded.  The GC carrier gas also passes
       through the 6-port valve and is returned to the GC.  With
       the 6-port valve in the Purge-Sorb position, the operation
       of the GC is in no way impaired; therefore, routine liquid
       injection analyses can be performed using the gas chroma-
       tograph in this mode.  After the sample has been purged, the
       6-port valve is turned to the desorb position (See Figure 4).
       In this configuration, the trap is coupled in series with
       the gas chromatographic column allowing the carrier gas to
       backflush the trapped materials into the analytical column.
       Just as the valve is actuated the power is turned on to the
       resistance wire wrapped around the trap.  The power is
       supplied by an electronic temperature controller.  A
       temperature sensor is attached to the silica gel/charcoal
       area of the trap and wrapped with a double layer of
       resistance wire.

-------
       The trap is rapidly heated to  180°C with minimal tempera-
       ture overshoot and maintained  at that temperature.   Under
       these conditions, the Tenax area of the trap, wrapped with  a
       single layer of resistance wire is maintained at
       approximately 130°C.  The trapped compounds are released
       as a "slug" to the gas chromatograph  by this heat and
       backflush step.  Normally, packed columns with theoretical
       efficiencies near 500 plates/foot under programmed
       temperature conditions can accept such desorb injections
       without altering peak geometry.  Substituting a
       non-controlled power supply, such as a manually operated
       variable transformer, will provide non-reproducible
       retention times and poor quantitative data unless the
       Injection Procedure (8.9.2) is used.
       NOTE:   It is acceptable to heat the Tenax area of the trap
       to 180°C.  However, this practice generally causes early
       trap failure.   Trap failure is characterized by a pressure
       drop in excess of three pounds per square inch across the
       trap during purging or by poor bromoform sensitivities.
4.1.4  Several Purge and Trap devices are now commercially  avail-
       able.   It is recommended that the following be taken into
       consideration if a unit is to be purchased:
       a. Be sure that the unit is completely compatible with the
          gas chromatograph to be used for the analysis.

-------
            b. Select a 5 mL purging device that meets the requirements
               shown in (4.1.1).
            c. The sorbant and sorbant areas of the trap must meet or
               exceed the specifications listed in (4.1.2).
            d. With the exception of sample introduction, select a unit
               that has as many automated purge trap functions as
               possible.
4.2. Gas chromatograph — The gas chromatograph column oven must be
     capable of operating at 40°C ± 1°C, as well as temperature
     programming.  The gas chromatograph must be equipped with automatic
     flow controllers so that the column flow rate will remain constant
     throughout the temperature program.  It may be necessary to cool
     the column oven down to less than 30°C.  See (8.9.2)  A
     subambient column temperature controller may be required to achieve
     this.
     4.2.1  Gas chromatographic columns — The gas chromatographic
            columns and conditions listed below are recommended for the
            analysis of complex mixtures of organohalides contained in
            water samples.  If these columns or conditions do not
            adequately resolve the organohalides,  the analyst may vary
            the liquid phase, loading level, solid support, mesh range
            column length, column internal diameter,  or operating
            temperatures in order to provide the required resolution.
            The analyst must have on file adequate precision and
            accuracy data to demonstrate that the  modified system
            provides data equivalent to that shown in Section 11 under
            single laboratory accuracy and precision.  Capillary columns

-------
and their associated specialized injection techniques are
not currently acceptable.
4.2.1.1 Column I is a highly efficient column which provides
        outstanding separations for a wide variety of
        organic compounds.  Because of its ability to
        resolve complex mixtures of organochlorine
        compounds, Column I should be used as the primary
        analytical column (See Figure 5).
        Column I parameters: — Dimensions: eight feet long
        x 0.1 inch ID stainless steel or glass tubing.
        Packing: 1% SP-1000 on Carbopack-8 (60/80) mesh.
        Carrier Gas: helium at 40 mL/minute.  Temperature
        program sequence:  45°C isothermal for three
        minutes, program at 8°C/minute to 220°C then
        hold for 15 minutes or until all compounds have
        eluted.
        NOTE:  It has been found that during handling,
        packing, and programming, active sites are exposed
        on the Carbopack-B packing.  This results in tailing
        peak geometry and poor resolution of many
        constituents.  To protect the analytical column,
        pack the first 5 cm of the column with 3% SP-1000 on
        Chromosorb-W 60/80 mesh followed by the Carbopack-B
        packing.  Condition the precolumn and the Carbopack
        columns with carrier gas flow at 220°C overnight.
        Pneumatic shocks and rough treatment of packed
        columns will cause excessive fracturing of the
        Carbopack.  If pressure in excess of 60 psi is

-------
               required to obtain 40 mL/minute carrier flow, the
               column should be repacked.
     4.2.1.2   Column II provides unique organohalide-separations
               when compared to those obtained from Column I (see
               Figure 6).  However, since the resolution between
               various compounds is generally not as good as those
               with Column I, it is recommended that Column II be
               used as a qualitative confirmatory column for
               unknown samples when GC/MS confirmation is not
               possible.
               Column II parameters — Dimensions:  six feet long x
               0.1 inch ID stainless steel or glass.  Packing:
               n-octane on Porisil-C (100/120 mesh).  Carrier Gas:
               helium at 40 cc/minute.  Temperature progranr
               sequence:  50°C isothermal for three minutes,
               program at 6°/minute to 170°C, then hold for
               four minutes or until all compounds have eluted.
4.2.2  Detector — A halogen specific detector must be used in
       order to eliminate misindentifications due to non organo-
       haTides which are coextracted during the purge step.
       a. A Hall model 700-A available from Tracor has been tested
          and found to provide the sensitivity needed to produce
          meaningful analyses down to 0.10 ug/L for most organo-
          halides with a relative standard deviation of less than
          10%.
          Recommended operating conditions for Hall 700-A Detector:
          Reactor tube:  nickel 1/16 inch O.D.

-------
   Reactor temperature:  810°C (Optimize the detector
   reactor temperature by analyzing a standard mixture
   containing 10 ng/uL of methylene chloride, chloroform,
   bromoform and tetrachloroethylene over a temperature
   range from 700°C through 1000°C.  Select the tempera-
   ture that provides maximum response to the compounds of
   highest interest.)
   Reactor base temperature:  250°C
   Electrolyte: 100% n-propyl alcohol
   Electrolyte flow rate:  0.8 ml/minute
   Reaction gas:  hydrogen at 40 mL/minute
   Carrier gas:  helium at 40 mL/minute
b. Other halogen specific detectors including other
   electrolytic conductivity systems and microcoulometric
   titration can be used.  The stability and sensitivity of
   these detectors normally limit the method to measurements
   down to 1.0 ug/L with a relative standard deviation near
   10%.
c. GC/MS techniques are acceptable options to halogen
   specific detectors for quantitative measurements.  The
   analyst must have on file single laboratory accuracy and
   precision data at 0.1 times the required maximum
   contaminant level to show that the system is adequately
   sensitive to be used for this application.  Additionally,
   recoveries should be within 20% of the true value with a
   relative standard deviation of less than 15% for repli-
   cate sample.

-------
                d. Flame Ionization and electron capture detectors are not
                   acceptable.  Approval for their use must be obtained
                   according to Section 3.1.
    4.3  Sample containers — 40-ml screw cap vials sealed with Teflon faced
         silicone septa.
    4.4  Syringes — 5 ml hypodermic with luerlok tip (2 each).
    4-.5  Micro syringes — 10, 100 uL.
    4.6  Micro syringe — 25 uL with a 2" by 0.006 inch I.D. needle
         (Hamilton #702N or equivalent).
    4.7  Syringe valve — 2-way with Luer ends (three each).
    4.8  Volumetric flasks — modified 500 and 1000-mL.   See Figure 7.
         (Special order from specialty glassware vendor.)
    4.9  Syringes — 0.5, 1.0, 5-mL gas-tight with shut off valve.
5.  Reagents and Materials
    5.1  Reagent safety precautions
         5.1.1  The toxicity or carcinogenicity of each reagent in this
                method have not been precisely defined;  however, each
                chemical compound should be treated as a potential health
                hazard.  From this viewpoint, exposure to these chemicals
                must be reduced to the lowest possible level by whatever
                means available.  The laboratory is responsible for
                maintaining a current awareness file of OSHA regulations
                regarding the safe handling of the chemicals specified in
                this method.  A reference file of material data handling
                sheets should also be made available to all personnel
                involved in the chemical analysis.

-------
     5.1.2  Vinyl chloride has been classified as a known human or
            mammalian carcinogen.  See Reference 1 for special
            precautions and permissible exposure limits for vinyl
            chloride.
5.2  Trap Materials
     5.2.1  Porous polymer packing — 60/80 mesh chromatographic grade
            Tenax GC (2,6-diphenyl-p-phenylene oxide).
     5.2.2  OV-1 (3X) on Chromosorb-W 60/80 mesh.
     5.2.3  Silica gel — 35/60 mesh Oavison, grade-15 or equivalent.
     5.2.4  Coconut charcoal — (26 mesh) Barnaby Chaney, CA-580-26 lot
            # M-2649 or equivalent.
5.3  Column packing — SP-1000 (1%) on Carbopack-8 (60/80 mesh)
     available from Supelco.
5.4  Column packing — n-Octane on Porasil-C (100/120 mesh) available
     from Waters Associates.
5.5  Column packing — SP-1000 (3%) on Chromosorb-W (60/80 mesh)
     available from Supelco.
5.6  Dechlorinating compound — crystalline sodium thiosulfate, A.C.S.
     Reagent Grade or sodium sulfite, A.C.S. reagent grade.
5.7  Activated carbon (for preparation of reagent water) --
     Filtrasorb-200 or Filtrasorb-400, available from Calgon Corp.,
     Pittsburgh, PA, or equivalent.
5.8  Reagent water
     5.8.1  Reagent water is defined as water free of interference for
            the compounds to be determined when employed in the purge
            and trap procedure described herein.  (See 9.11)   It  is

-------
            generated by passing distilled or tap water through a carbon
            filter bed containing about one pound of activated carbon.
     5.8.2  A Millipore Super-Q Water System or its equivalent may be
            used to generate reagent water.
            NOTE:  1,2-dichloroethane and methylene chloride are common
            contaminants found in depleted Super-Q Water Systems.
     5.8.3  Reagent water may also be prepared by boiling water for 15
            minutes. Subsequently, while maintaining the temperature at
            90°C, bubble a contaminant free inert gas through the
            water for one hour.  While still hot, transfer the water to
            a narrow mouth screw cap bottle with a Teflon seal.
5.9  Standards
     5.9.1  Obtain at least 97% pure reagent grade reference standards.
            From commercial sources or, as they become available, from
            the EMSL-CI Quality Assurance Branch Repository, Cincinnati,
            Ohio 45268.
     5.9.2  Vinyl chloride — 99.9% Pure vinyl chloride in cylinders is
            available from Ideal Gas Products, Inc., Edison, New Jersey
            and upon request from Matheson, East Rutherford, New Jersey.
            (See Reference 1 for safety precautions.)  Certified
            mixtures of vinyl chloride in nitrogen at 1.0 and 10.0 ppm
            are available from several sources.
5.10 Standard Stock Solutions (compounds boiling above room temperature)
     5.10.1 Place about 9.8 ml methyl alcohol into a 10-mL ground glass
            stoppered volumetric flask.
     5.10.2 Allow the flask to stand unstoppered for about 10 minutes or
            until all alcohol wetted surfaces have dried.

-------
     5.10.3 Weigh the flask to the nearest 0.1 mg.
     5.10.4 Using a 100-uL syringe, immediately add 2 drops of the
            reference standard to the flask, then reweigh.  Be sure that
            the 2 drops fall directly into the alcohol without
            contacting the neck of the flask.
     5.10.5 Dilute to volume, stopper, then mix by inverting the flask
            several times.
     5.10.6 Calculate the concentration in micrograms per microliter
            from the net gain in weight.
     5.10.7 Transfer the standard solution to a 10-mL screw-cap bottle
            with a Teflon cap liner.
     5.10.8 Store the solution at 4°C.
            NOTE:  With the exception of 2-chloroethylvinyl ether,
            standard solutions prepared in methyl alcohol are stable up
            to four weeks when stored under these conditions.  They
            should be discarded after that time has elapsed.  Standard
            solutions containing 2-chloroethylvinyl ether are stable for
            one week.
5.11 Standard Stock Solutions (Gaseous Compounds)
     5.11.1 Place about 9.8 mL of methyl alcohol  into a 10.0-mL ground
            glass stoppered volumetric flask.
     5.11.2 Allow the flask to stand  unstoppered  about 10 minutes or
            until all alcohol wetted  surfaces have dried.
     5.11.3 Weigh to the nearest 0.1  mg.
     5.11.4 Fill a 5-mL valved gas-tight syringe  with the reference
            standard to the 5.0 ml mark.
     5.11.5 Lower the needle to 5 mm  above the methyl alcohol menicus.

-------
     5.11.6 Slowly inject the reference standard into the neck of flask
            (the gas will rapidly dissolve into the methyl alcohol).
     5.11.7 Immediately reweigh the flask to the nearest 0.1 mg.
     5.11.8 Dilute to volume, stopper, then mix by inverting the flask
            several times.
     5.11.9 Transfer the standard solution to a 10-mL screw cap bottle
            with a Teflon cap liner.
     5.11.10   Store stock solutions at < 0° C.
     5.11.11   Stock solutions prepared from gaseous compounds are not
               stable for periods exceeding one week.  They should be
               discarded after that time.
     5.11.13   Calculate the concentration in micrograms per micro!iter
               from the net gain in weight.
5.12 Calibration Standards
     5.12.1 In order to prepare accurate aqueous standard solutions, the
            following precautions must be observed.
            a. Do not inject more than 20 uL of alcoholic standards into
               100 ml of reagent water.
            b. Use a 25-uL Hamilton 702N microsyringe or equivalent.
               (Variations in needle geometry will adversely effect the
               ability to deliver reproducible volumes of methanolic
               standards into water.)
            c. Rapidly inject the alcoholic standard into the expanded
               area of the filled volumetric flask.  (Figure 7.)  Remove
               the needle as fast as possible after injection.
The validity of the entire standard preparation scheme is monitored on a
quarterly basis by analyzing Certified Reference Standards as they
become available (See 9.6).

-------
            d.  Mix aqueous standards by inverting the flask three times
               only.
            e.  For standards prepared in 500 or 1000-mL flasks, discard
               the contents contained in the neck of the flask.  Fill
               the sample syringe from the standard solution contained
               in the expanded area of the flask as directed in
               paragraph 8.5.
            f.  Never  use pipets to dilute or transfer aqueous standards
               and samples.
            g.  Aqueous standards are not stable and should be discarded
               after  one hour unless stored and sealed according to
               6.4.8  and 6.4.9.6.
     5.12.2 Prepare,  from the standard stock solutions, secondary
            dilution  mixtures in methyl alcohol so that a 20 uL
            injection into 100, 500, or 1000 ml of reagent water will
            generate  a calibration standard which produces a response
            close (±20%) to that of the unknowns.
            Note:  Comrnercially available primary or secondary dilutions
            of organohalides in methane! may be used only if they are
            supplied  with a certified analysis.
     5.12.3 Purge and analyze the aqueous calibration standards in the
            same manner as the unknowns.
5.13 Calibration using certified mixtures of vinyl chloride in nitrogen.
     5.13.1 The gas chromatograph can be calibrated for vinyl chloride
            by injecting a known volume of a certified mixture of vinyl
            chloride in nitrogen into the purging device as other
            aqueous standards are being purged.

-------
     5.13.2 Fill  the purging device with 5.0 ml reagent water or aqueous
            calibration mixture.
     5.13.3 Start to purge the aqueous mixture.  Inject a known volume
            (between 100 and 2000 uL) of the calibration gas with a gas
            tight syringe directly into the purging device.  Slowly
            inject the gaseous sample through the septum seal at the top
            of the purging device at ZOOOuL/minute.  Do not inject the
            standard through the aqueous sample inlet needle.  Inject
            the gaseous standard before five minutes of the eleven
            minute purge time have elapsed.
     5.13.4 Record the volume of gaseous standard (ml), the barometric
            pressure (mm of mercury) and the temperature of the vinyl
            chloride gas (degree C) at the time of standardization.
5.14 Laboratory Control Standard (LC) (0.40 ug/L)
     5.14.1 From the standard stock solutions, prepare a secondary
            dilution in methyl alcohol containing 10 ng/uL of each
            compound normally monitored.
            NOTE:  It may be necessary to prepare two or more LC so that
            all of the compounds in each mixture are adequately resolved
            for quantitative measurement.
     5.14.2 Inject 20.0 uL of this mixture daily into 500 mL of reagent
            water and analyze according to the Procedure, Section 8.
5.15 Certified Reference Standard — Obtain Certified Reference Standard
     mixtures from the Environmental Monitoring and Support Laboratory -
     Cincinnati,  Quality Assurance Branch as they become available.
     Dilute in reagent water according to the  instructions supplied with
     the mixture.

-------
6.  Sample Collection and Handling
    6.1  The sample containers should have a total volume in excess of 20 ml.
         6.1.1  Narrow mouth screw cap bottles with the TFE fluorocarbon
                faced silicone septa cap liners are strongly recommended.
                Crimp-seal serum vials with TFE fluorocarbon faced septa are
                acceptable if the seal is properly made and maintained
                during shipment and storage.
    6.2  Sample Bottle Preparation
         6.2.1  Wash all sample bottles and TFE seals in detergent solution.
                Rinse with tap water and finally with distilled water.
         6.2.2  Allow the bottles and seals to air dry at room temperature,
                place them in a 105°C oven for one hour, and allow to cool
                in an area known to be free of organics.
                NOTE:  Do not heat the TFE seals for extended periods of
                time, that is, for more than one hour because at 105°C the
                silicone layer slowly begins to shrink.
         6.2.3  When cool, seal the bottles with the TFE seals that will be
                used for sealing the samples.
    6.3  Sample Preservation — Either sodium thiosulfate or sodium sulfite,
         chemical  dechlorinating agents, is added to samples containing free
         chlorine  in order to arrest the formation of trihalomethanes after
                           2
         sample collection.    If chemical preservation is employed, the
         preservative is also added to the blanks.  The crystalline chemical
         preservative (2.5 to 3 mg/40 mL) is added to the empty sample
         bottles just prior  to shipping to the sampling site.  See Table II
         to determine the stability of various organohalides in the presence
         of sodium thiosulfate or sodium sulfite.

-------
6.4  Sample Collection
     6.4.1  Collect a minimum of two samples from each sample source,
            field duplicate-! (FD-1) and field duplicate-2 (FD-2).
     6.4.2  Fill the sample bottles in such a manner that no air bubbles
            pass through the sample as the bottle is being filled.
     6.4.3  Seal the bottles so that no air bubbles are entrapped inside
            them.
     6.4.4  Maintain the hermetic seal on the sample bottle until time
            of analysis.
     6.4.5  Sampling from a water tap.
            6.4.5.1 Turn on water and allow the system to flush.  When
                    the temperature of the water has stabilized, adjust
                    the flow to about 500 mL/minute and collect FD-1 and
                    FD-2 samples from the flowing stream.
     6.4.6  Sampling from an open body of water.
            6.4.6.1 Fill a 1 quart wide mouth bottle or 1 liter beaker
                    with sample from a representative area.  Carefully
                    fill FD-1 and FD-2 from the sampling container as
                    noted in 6.4.1 through 6.4.4.
     6.4.7  If preservative has been added to the sample bottles, then
            fill with sample just to overflowing, seal the bottle, and
            shake vigorously for one minute.
     6.4.8  Sealing practice for septum seal screw cap bottles.
            6.4.8.1 Open the bottle and fill to overflowing, place on a
                    level surface, position the TFE side of the septum
                    seal upon the convex sample meniscus and seal the
                    bottle by screwing the cap on tightly.

-------
       6.4.8.2 Invert the sample and lightly tap the cap on  a  solid
               surface.  The absence of entrapped air  indicates  a
               successful seal.  If bubbles are present, open  the
               bottle, add a few additional drops of sample  and
               reseal bottle as above.
               NOTE:  If the septum seals are inverted, that is, if
               the silicone side comes in contact with the sample
               significant organohalide losses will occur in
               shipment and storage.
6.4.9  Blanks
       6.4.9.1 Field Reagent Blanks (FRB) must be prepared and
               accompany the samples wherever the samples are
               shipped or stored.  If the samples are  immediately
               analyzed at the sampling site, FRB are  not required.
       6.4.9.2 Prepare FRB in duplicate at the laboratory by
               filling and sealing a minimum of two sample bottles
               with pre-tested reagent water just prior to shipping
               the sample bottles to the sampling site.
       6.4.9.3 If the sample is to be preserved, add an identical
               amount of preservative to the FRB.
       6.4.9.4 Ship the FRB to and from the sampling site along
               with the sample bottles.
       6.4.9.5 Store the FRB and the samples collected from  a  given
               source, called a sample set, together.  A sample  set
               is defined as all the samples collected from  a  given
               source at a given time, for example, at a water

-------
                        treatment plant, the duplicate raw source waters,
                        the duplicate finished waters and the duplicate FRB
                        samples comprise the sample set.
                6.4.9.6 Store the sample set at 4°C in an area known to be
                        free of organic vapors.  The maximum recommended
                        holding times are:  vinyl chloride - six days,
                        carbon tetrachloride - 27 days, 1,2-dichloroethane -
                        27 days, 1,1,1-trichloroethane - 21 days, 1,1,2-tri-
                        chloroethylene - 27 days and 1,1,2,2-tetra-
                        chloroethylene - 27 days.  See Table II for maximum
                        storage time for other organohalides.
7.   Conditioning Traps
    7.1   Condition newly packed traps overnight at 180°C by backflushing
                                        •
         with an inert gas flow of at least 20 ml_/minute.
         7.1.1  Vent the trap effluent to the room, not to the analytical
                column.
    7.2  Daily prior to use, condition traps 10 minutes while backflushing
         at 180°C.
         7.2.1  The trap may be vented to the analytical column; however,
                after conditioning the column must be programmed prior to
                analysis of samples.
8.   Procedure
    8.1   Adjust the purge gas (nitrogen or helium) flow rate to 40 mL/minute.
    8.2  Attach the trap inlet to the purging device.  Turn the valve to the
         purge-sorb position (Figure 3).
    8.3  Open the syringe valve located on the purging device sample
         introduction needle.

-------
8.4  Remove the plungers from two 5-mL syringes  and  attach  a  closed
     syringe valve to each.
8.5  Warm the sample to room temperature then open the FD-1 bottle {or
     standard) and carefully pour the sample into one of the  syringe
     barrels until it overflows.  Replace the syringe plunger and
     compress the sample.  Open the syringe valve and vent  any residual
     air while adjusting the sample volume to 5.0 ml.  Close  the valve.
8.6  Fill the second syringe in an identical manner  from the  FD-1 sample
     bottle.  This second syringe is reserved for a  duplicate FD-1
     analysis, if necessary.
8.7  Attach the syringe-valve assembly to the syringe valve on the
     purging device.
8.8  Be sure that the trap  is cooler than 25°C then open the syringe
     valve and inject the sample into the purging chamber.  Close both
     valves.  Purge the sample for 11.0 ± 0.1 minutes.
8.9  After the 11 minute purge time, attach the trap to the chroma-
     tograph (turn the valve to the desorb position) and introduce the
     trapped materials to the GC column by rapidly heating the trap to
     180°C while backflushing the trap with an inert gas between 20
     and 60 mL/minute for four minutes.
     8.9.1  If the trap can be rapidly heated to 180°C and maintained
            at this temperature, the GC analysis can begin as the sample
            is desorbed, that is, while the column is at the  initial
            45°C operating temperature.  The equipment describee! in
            Figure 4 will perform accordingly.
     8.9.2  With other types of equipment (see 4.1.4) where the trap is
            not rapidly heated or is not heated in a reproducible

-------
            manner, it is necessary to transfer the contents of the trap
            into the analytical column at  30°C where it is once again
            trapped.  Once the transfer is complete, which takes four
            minutes, the column is rapidly heated to the initial
            operating temperature for analysis.
            NOTE:  In some cases, it may be necessary to cool the
            analytical column to 0°C.
     8.9.3  If injection procedure 8.9.1 is used and the early eluting
            peaks in the resulting chromatogram have poor geometry or
            variable retention times, then method 8.9.2 must be used.
8.10 While the extracted sample is being introduced into the gas
     chromatograph, empty the purging device using the sample intro-
     duction syringe, follow by two 5-mL flushes of reagent water.
     After the purging device has been emptied, leave the syringe valve
     open to allow the purge gas to vent through the sample introduction
     needle.
8.11 After desorbing the sample for approximately four minutes,
     recondition the trap by returning the valve to the sorb position.
     Wait 15 seconds, then close the syringe valve on the purging
     device, allowing the purge gas to pass through the trap.  Maintain
     the trap temperature at 180°C.  After approximately seven
     minutes, turn off the trap power and open the syringe valve.
     NOTE:  If the operations described in 8.11 are omitted, large
     amounts of water will be injected into the column.  For certain
     detectors, this will cause numerous large narrow peaks or detector
     response quenching to occur in the early elution area of the
     chromatogram on subsequent analysis.

-------
    8.12 Analyze each FD-1 and FRB from the sample set in an identical
         manner (see 6.4.9.5) on the same day.
    8.13 Prepare single point calibration standards from the standard stock
         solutions in reagent water that are close (±20%) to the unknown in
         composition and concentration (9.5).  The concentrations should be
         such that no more than 20 uL of the secondary dilution need be
         added to 100 to 1000 ml reagent water to produce a standard at the
         same level as the unknown.
    8.14 As an alternative to Section 8.13, prepare a calibration curve for
         each organohalide containing at least three points, two of which
         must bracket the unknown.  Check the validity of calibration curves
         daily by analyzing the LC (9.5).
    8.15 As a second alternative to Section 8.13, internal standard
         calibration techniques may be used.  The following organohalides
         are recommended for this purpose:   Bromochloromethane,
         2-bromo-l-chloropropane or 1,4-dichlorobutane.  The internal
         standard is added to the sample just before purging.  Check the
         validity of the internal standard  calibration factors daily by
         analyzing the LC sample (9.5).
    8.16 For gaseous vinyl chloride, prepare single point calibration
         standards within 20X of the sample or a calibration curve according
         to 8.14.  Check the validity daily by analyzing a gaseous standard.
9.  Quality Control
    9.1  Each laboratory that uses this method is required to operate a
         formal  quality control program.  The minimum requirements of this
         program consist of an initial demonstration of laboratory
         capability and the analysis of spiked samples as a continuing check

-------
     on performance.  The laboratory is required to maintain performance
     records to define the quality of data that is generated.  As Agency
     sponsored interlaboratory data is gathered for each parameter,
     ongoing performance checks must be compared with these performance
     criteria to determine if the results of analyses are within
     accuracy and precision limits expected of the method.
     9.1.1  Before performing any analyses, the analyst must demonstrate
            the ability to generate acceptable accuracy and precision
            with this method.  This ability is established as described
            in Section 9.2.
     9.1.2  In recognition of the rapid advances that are occurring in
            chromatography, the analyst is permitted certain options to
            improve the separations or lower the cost of measurements.
            Each time such modifications are made to the method, the
            analyst is required to repeat the procedure in Section 9.2.
     9.1.3  The laboratory must spike and analyze a minimum of 10% of
            all FD-2 samples to monitor continuing laboratory
            performance.  This procedure is described in Section 9.4.
9.2  To establish the ability to generate acceptable accuracy and
     precision in the use of this method, the analyst must perform the
     following operations.
     9.2.1  For each parameter normally measured,  spike reagent water
            with a methanolic standard solution so that the resulting
            aqueous mixture contains each compound near the maximum
            contaminant level.
     9.2.2  Analyze a minimum of four different spiked samples.  A
            representative natural water may be used in place of the

-------
            reagent water, but one or more additional aliquots must be
            analyzed to determine background levels, and the spike level
            must exceed twice the background level for the test to be
            valid.  Analyze the aliquots according to Section 8.
     9.2.3  Calculate the average percent recovery, (R), and the
            relative standard deviation of the concentration found.
            Natural water background corrections must be made before R
            calculations are performed.
     9.2.4  After the Agency collects interlaboratory values for R and
            s, then the laboratory must meet method performance criteria
            established from these data.
9.3  The analyst must calculate method performance criteria and define
     the performance of the laboratory for each spike concentration and
     parameter being measured.
     9.3.1  Calculate upper and lower control limits for method
            performance:
               Upper Control Limit (UCL)  =  R + 3 s
               Lower Control Limit (LCL)  =  R - 3 s
            where R and s are calculated as in Section 9.2.3.
            The UCL and LCL can be used to construct control charts
     *
            that are useful in observing trends in performance.
     9.3.2  The laboratory must develop and maintain separate accuracy
            statements of laboratory performance for water samples.  An
            accuracy statement of the method is defined as R ± s.  The
            accuracy statement should be developed by the analysis of
            four aliquots of natural water as described in Section

-------
            9.2.2, followed by the calculation of R and s.  Alternately,
            the analyst may use four natural water data points gathered
            through the requirement for continuing quality control in
            Section 9.4.  The accuracy statments should be updated
            regularly.
9.4  The laboratory is required to collect a portion of their samples in
     duplicate to monitor spike recoveries.  The frequency of spiked
     sample analysis must be at least 10% of all samples or one sample
     per month, whichever is greater.   One aliquot of the sample must
     be spiked and analyzed as described in Section 9.2.  If the
     recovery for a particular parameter does not fall within the
     control limits for method performance, the results reported for the
     parameter in all samples processed as part of the same set must be
     qualified as described in Section 10.4.  The laboratory should
     monitoring the frequency of data so qualified to ensure that it
     remains at or below 5%.
9.5  Analyze the 0.40 ug/L LC sample daily before any samples are
     analyzed.  Instrument status checks are obtained from these data.
     In addition, response factor data obtained from the 0.40 ug/L LC
     standard can be used to estimate the concentration of the unknowns.
     From this information the appropriate single point standard
     dilutions can be determined for 8.13.
     9.5.1   Calculate the response factor for each compound contained in
            the LC sample.  Compare the values to the mean of previously
            determined factors.  If they are different by more than
            ±10%, then run a duplicate LC sample to insure that the
            instrument is operating properly.

-------
            NOTE:  Generally, for electrolytic conductivity detectors,
            the first analysis of the day produces low response factors.
     9.5.2  If Calibration Procedure 8.14 or 8.15 is used, then calcu-
            late the concentration of each component in the 1C sample.
            Each value must be within ±10% of the true value.  If they
            are not, then analyze a duplicate LC sample.  If the
            value(s) is still off by more than ±10%, then prepare a new
            calibration curve.
9.6  Analyze the Certified Reference Standard (5.15) on a quarterly
     basis.  Dilute with reagent water according to the instructions
     supplied with the mixture.  The resulting analysis must agree
     within (± 10%) of the certified values.  If they do not, then
     evaluate each step of the calibration procedure to identify and
     correct the source of error.  Analyze the Certified Reference
     Standard whenever there is a reason to suspect the validity of
     laboratory generated standard dilutions.
9.7  Analyze the FRB or a LRB to monitor for potential  interferences as
     described in Sections 3.2, 3.3 and 3.6.
9.8  Perform the following instrument status checks daily, using the
     data gathered from standards, duplicate analyses and the LC sample.
     9.8.1  Peak Geometry Check
            9.8.1.1 All of the peaks contained in the LC chromatogram
                    must appear to be sharp and symmetrical.  Peak
                    tailing in excess of that shown in  the method
                    chromatogram Figure 5 must be corrected.  Tailing
                    problems are generally traceable to:

-------
               a.   Active sites on GC column - repack.
               b.   Reactor temperature too low - see 4.2.2 (Hall
                   detector).
               c.   Reactor base temperature too low - see 4.2.2
                   (Hall  detector).
               d.   Contaminated reactor tube-recondition or replace
                   (Hall  detector).
               e.   Contaminated reactor transfer line - wash with
                   concentrated HC1 then dry or replace (Hall
                   detector).
               f.   Detector electrolyte flow too low - see 4.2.2.
               g.   Spent  ion exchange column - Replace (Hall
                   detector).
               h.   Reactive area or large dead volume between
                   column and detector.  Shorten transfer lines.
                   Recondition or replace transfer lines and
                   reactor base.
       9.8.1.2 If  only the compounds eluting before chloroform give
               random responses, unusually wide peak widths, are
               poorly resolved, or are missing, the problem is_
               usually traceable to the trap/desorber.  See
               Sections 8.9.3, 4.1.2.
       9.8.1.3 If  only brominated compounds show poor peak geometry
               or  do not  properly respond at low concentrations,
               repack the trap.  Excessive Hall detector reactor
               temperatures can also cause low bromoform response.
       9.8.1.4 If  negative peaks appear in the chromatogram,
               replace the ion exchange column and replace
               electrolyte (Hall detector).
9.8.2  Randomly select and analyze 10% of all FD-2 samples.  Field
       duplicate-! and FD-2 analyses should agree within 10%.

-------
                Poor agreement is  generally traceable to:
                a.  Pneumatic  leaks,  especially around the  purging device or
                   the  Hall detector reactor inlet and exit.
                b.  Electronic problems.
                c.  Inexperienced operator.
                d.  Sampling and storage  problems.
    9.10  Maintain  a record of the  retention times  for each organohalide
         using  data gathered  from  LC samples and standards.
             9.10.1  Calculate the  average retention time daily for each
                     compound normally  monitored  and the  variance
                     encountered  for the analyses.
             9.10.2 If individual  retention times  vary by more than 10%
                     over an eight  hour period or does not fall within 1056
                     of an established  norm, the  system is "out of
                     control." The  source of retention data variation must
                     be corrected before acceptable data can be generated.
    9.11  The FRB and LRB analysis  should represent less than a 0.01 ug/L
         response  or less than a 10% interference  for those compounds that
         are reported.
    9.12  Any instrument not  performing according to 9.5, 9.6 and 9.10
         specifications should be  considered "out  of control."  The
         instrument must be  "in control" before acceptable data can be
         generated.
10.  Calculations
    10.1  Identify each  organohalide  in the sample chromatogram by comparing
         the retention  time  of the suspect peak to the data gathered in

-------
         9.10.  The retention time of the suspect peak must fall within the
         limits established in 9.10 for single column identification.
    10.2 Quantify the unknowns by comparing the peak area or peak height of
         the unknowns to the standard peak area or height.  Round off the
         data to the nearest .01 ug/L or two significant figures.
           ,.  a  peak height sample    x (concn. std, ug/L)
         ug/     peak height standard
    10.3 Calibration using vinyl chloride in nitrogen standards.
         ug/L  -  7^    (C)  (V)  (152)
                       T + 273
         where ug/L = aqueous equivalent concentration of standard
              P = atmospheric pressure (mm)
              C s concentration of gaseous standard in ppm (volume:volume).
              V « volume of standard injected into purging device (mL)
              T - temperature of gaseous standard (degrees C.)
    10.4 For samples processed as part of a set when laboratory spiked
         sample recovery falls outside of the control limits in Section 9.4,
         label the affected paratmeter as suspect.
11.  Accuracy and Precision and Method Detection Limits
    11.1 Precision and Accuracy for the Purge and Trap Method Using  the Hall
         700-A Detector under the conditons described in Section 4.2.2.
         11.1.1 Both Ohio River water (ORW) and carbon filtered tap  (CFT)
                water were spiked with known amounts of organohalides.  The
                spiked solutions were then sealed in septum-seal vials both
                with and without sodium thiosulfate (thio) and  sodium
                sulfite (sulfite) then stored on the bench top  for up to

-------
            four weeks.  Samples were randomly analyzed on four
            occasions.  When matrix effects were noted or suspected,
            these data were not included in the following single
            laboratory precision and accuracy statement.  Table II shows
            the accuracy and precision data obtained from this study.
11.2 Precision and Accuracy for Purge and Trap Method using Hall 700
     Detector.
     11.2.1 Organic-free water was spiked with mixtures of
            organohalides.  The spiked water was used to fill septum seal
            vials which were stored under ambient conditions.  The
            spiked samples  were randomly analyzed over two weeks.  The
            data listed in  Table III lists the accuracy and precision
            data obtained from this Study.
11.3 Method detection Limits (MDL) were determined using the Hall 700-A
     detector.  Carbon filtered tap water was spiked with a mixture of
     organihalides.   Seven  sample bottles were filed from the mixture
     then analyzed.   Spike  levels were adjusted to provide a response at
     2 to 10 times the noise level at the 2X attenuation setting.  The
     resulting data appear  in Table IV.

-------
                     TABLE  I
OR6ANOHALIDES TESTED USING PURGE AND TRAP METHOD
Compound

chloromethane
bromomethane
dichlorodifloromethane
vinyl chloride
chloroethane
di chloromethane
fluorotri chloromethane
allylchloride
1,1-dichloroethylene
bromoch 1 oromet hane
1,1 -di chloroethane
trans-1 ,2-dichloro-ethylene
cis-1 ,2-dichloro-ethylene
chloroform
1,2-dichloroethane
di bromomethane
1,1,1 -tri chloroethane
carbon tetrachloride
dichloroacetonitrile
bromodi ch 1 oromethane
2,3-dichloropropene
1 ,2-dichloropropane
1,1-dichloropropene
trans-1, 3 dichloropropene
1,1,2-trichloroethylene
1 ,3-dichloropropane
chlorodibromomethane
1 , 1 ,2-tri chloroethane
cis-l,3-dichloropropene
1,2-dibromoethane
2-chloroethyl ethyl ether
2-chloroethyl vinyl ether
bromoform
1,1,1 ,2-tetrach 1 oroethane
1,2,3-trichloropropane
chlorocyclohexane
1,1, 2, 2-tetrachl oroethane
1 ,1 ,2,2-tetrachloroethylene
Retention Time
(SEC)
Column I Column II

90
130
157
160
200
315
431
475
476
509
558
605
605
641
684
698
756
781
884
819
891
895
904
913
948
973
989
991
992
1046
1056
1080
1154
1163
1279
1283
1297
1300

317
•423
A
317
521
607
A
A
463
760
754
563
726
725
921
895
786
664
A
877
A
997
A
997
787
A
997
1084
1078
1131
A
A
1150
1302
A
A
A
898
CAS Number0

74-87-39
74-83-9
75-71-8
75-01-4
75-00-3
75-09-2
75-69-4
107-05-1
75-35-4
74-97-5
75-34-3
156-60-5
156-59-2
67-66-3
107-06-2
74-95-3
71-55-6
56-23-57
3018-12-0
75-27-4
78-88-6
78-87-5
563-58-6
10061-02-6
79-01-6
142-28-9
124-48-1 •
79-00-5
10061-01-5
106-93-5
7081-78-9
110-75-8
75-25-2
630-20-6
96-18-4
542-18-7
79-34-5
127-18-4
Purging
Efficiency
Percent
91
85
B
101
90
76
96
B
97
88
89
94
92
88
95
98
B
94
87
10
92
" 91
92
B
90
89
B
87
88
85
64
54
19
71
89
B
B
58
88

-------
                                TABLE  I  (Continued)
Compound
pentachloroethane
1 -ch 1 orocyc 1 ohexene- 1
chl orobenzene
1-chlorohexane
bis-2-chloroethyl ether
1 ,2-dibromo-3-chloropropane
bromobenzene
o-chlorotoluene
bis-2-chloroisopropyl ether
m-di chl orobenzene
o-di chl orobenzene
p-di chl orobenzene
Retention Time
(SEC)
Column I Column II
1300
1345
1451
1499
1500
1560
1626
1927
1931
2042
2094
2127
A
1186
1130
1229
A
A
A
1320
A
1346
1411
1340
CAS Number^
76-01-7
930-66-5
108-90-7
544-10-5
111-44-4
96-12-8
108-86-1
95-49-8
108-60-1
541-73-1
95-50-1
106-46-7
Purging
Efficiency
Percent
B
B
83
8
B
9
B
B
B
68
B
. 70
A - Not Determined
B = Not Determined
C = Chemical Abstracts Service Registry Number

-------
                         TABLE II

Single Lab Accuracy and  Precision for  Purge  and  Trap Method
       Hall 700A Electrolytic Conductivity Detector
Compound
Chloromethane
bromomethane
vinyl chloride
di chl or odifluorome thane
chloroethane
di chloromethane
fluorotrichloromethane
1 , 1-dichloroethylene
allylchloride
bromochloromethane
1,1-dichloroethane
c1s+trans-l,2-dichloro-
ethylene
cis-l,2-d1chloro-
ethylene
1,2-dichloroethane
di bromomethane
1 , 1 , 1-trichloroethane
carbon tetrachlorlde
bromodichloromethane
2,3-dichloropropene
1 ,2-dichloropropane
1,1-dichloropropene
trans-1 ,3-dichloropro-
pene
cis-l,3-dichloropro-
pene
1 ,1 ,2-trichloroethane
1,3-dichloropropane
Spike
Level
0.40
0.40
0.20
0.40
0.40
0.20
0.40
0.40
0.40
0.40
0.20

0.20

0.40
0.20
0.40
0.40
0.20
0.20
0.20
0.40
0.40

0.40

0.40
0.40
0.40
Preserved
Sample
Thlo. Mean % Number of
C.F.TD O.R.WE Sulflte Recovery^ Samples
G
G,C
F
G
G
F
G
G
G,C
G
F

F

G
F
G
G
F
F
F
G
G

G,C

G,C
G
G
G
G
G
G
G
G
G
G
G,C
G
G

G

G
G
G
G
G
G
6.C
G
G

G,C '

G,C
G
G
G,A,B
G
F
G
G
F
G
G,A
G,C
G
F

F

G
F
G
G
F
F
F
G
G,A

G,C

G,C
G
G
93
85
no
103
93
85
90
88
85
90
95

95

88
110
100
93
90
100
95
95
88

88

90
95
98
16
8
12
12
20
17
21
18
8
19
17

17

20
17
5
20
17
17
14
20
18

4

4
15
21
Std.
Devia-
tion
.034
.025
.029
.081
.071
.024
.037
.037
.046
.038
.012

.011

.028
.014
.032
.032
.014
.013
.012
.014
.037

.000

.050
.024
.026
Rel. Maximum
Std. Holding
Devi a- Time
tion (Days)
8.5
6.3
15
20
18
12
9.3
9.3
12
9.5
6.0

5.5

7.0
7.0
8.0
8.0
7.0
6.5
6.0
3.5
9.3

.000

12.5
6.0
6.5
21
2
6
27
21
27
27
27
2
21
27

27

21
27
21
21
27
27
6
21
27

1

1
27
27

-------
TABLE II (Continued)
Spike
Level
Compound
ch 1 or od 1 bromomethane
1 , 1 ,2-trichloroethylene
1,2-dibromoethane
2-chloroethylether ether
2-chloroethylvlnyl ether
bromoform
1,1,1,2-tetrachloro-
e thane
1,2,3-trichloropropane
chlorocyclohexane
1,1,2,2-tetrachloro-
ethane
1,1,2,2-tetrachloro-
ethylene
pentachloroethane
1-chlorocyclohexene-l
chlorobenzene
1-chlorohexane
bis-2-chloroethylether
bromobenzene
o-chlorotoluene
bls-2-chlorolsopropyl
ether
m-dichlorobenzene
o-dichlorobenzene
p-dichlorobenzene
A - Matrix effect noted due
B - Matrix effect noted due
C - Matrix effect noted due
D - C.F.T. - carbon filtered
E - O.R.W - Ohio River water
F - Not determined.
ug/
0.
0.
0.
0.
0.
0.

0.
0.
0.

0.

0.
0.
0.
0.
0.
0.
0.
0.

0.
0.
0.
0.
to
to
to
20
20
40
40
40
20

40
40
40

40

20
40
40
40
40
40
40
40

40
40
40
40
the
the
the
C.F.TD
F
F
6
G
G
F

G
G
G

G

F
G
G
G
G,C
G,C
G
G

G
G
G
G
presence
presence
Preserved
Sample
Thio.
O.R.WE Sulfite
G
G
G
G
G
G

G
G
G

G

G
G
G
G
G,C
G,C
G
G

G
G
G
G
F
F
G,B
G,A
G
F

G
G
G

G,A

F
G
G
G
G,C
G,C
G
G

G
G
G
G
of sodium sulfite
Mean %
Std.
Number of Devia-
Recovery Samples
95
94
93
95
100
95

93
100
93

95

90
98
« 93
88
83
100
93
85

125
95
95
90
(use of
of sodiumthiosulfate (use
sample storage
(recommended
storage
17
17
18
18
21
17

20
20
21

18

17
21
21
18
4
16
20
20

21
21
21
20
reagent not
of reagent
time noted
tion
.014
.012
.050
.030
.031
.030

.032
.038
.033

.036

.019
.039
.051
.037
.022
.065
.047
.037

.11
.033
.053
.051
Rel. Maximum
Std. Holding
Devia- Time
tion (Daysl
7.
6.
12.
7.
7.
15.

8.
9.
8.

9.

9.
9.
12.
9.
5.
16.
12.
9.

28.
8.
13.
13.
0
0
5
5
8
0

0
5
3

0

5
8
8
3
5


3


3


27
27
21
27
27
27

21
21
27

21

27
27
27
21
1
9
21
21

27
27
27
21
recommended).
not recommended).
in maximum
tap water.




G - These data included in recovery




and standard








age



column)



•



deviation calculations.

-------
                          Table  III
Single Laboratory Accuracy and Precision for Trihalomethanes
        Hall 700  Electrolytic Conductivity  Detector
          Dose
         (ug/L)

            1.19
           11.9
          119
          Dose
         (ug/L)

            1.60
           16.0
          160
          Dose
         (ug/L)

            1.96
           19.6
          196
          Dose
          (ug/L)

            2.31
           23.1
          231
     Chloroform

     Number         Mean      Standard
     samples        (ug/L)    deviation

       12             1.21      0.14
        8            11.3       0.16
       11           105         7.9

Bromodi ch1oromethane
Number
samples
12
8
n
Mean
(ug/L)
1.52
15.1
145
Standard
deviation
0.05
0.39
10.2
                    Chlorodlbromomethane
Number
samples
12
8
11
Mean
(ug/L)
1.91
19.1
185
Standard
deviation
0.09
0.70
10.6
     Bromoform

     Number
     samples

       12
        8
       11
Mean
(ug/L)

  2.33
 22.5
223
Standard
deviation

  0.16
  1.38
 16.3

-------
                                     TABLE  IV

                METHOD DETECTION LIMITS FOR SELECTED ORGANOHALIDES
Relative
Spike Average Standard
Level Recovery Deviation
Compound
Methyl chloride
Vinyl chloride
Methyl bromide
Ethyl chloride
1 , 1 -0 i ch 1 oroethy 1 ene
1,1-Oichloroethane
Methyl ene chloride*-
cis+trans-l,2-0ichloroethylene
Chloroform
1,2-Oichloroethane
1 , 1 , 1-Trichloroethane
Carbon tetrachloride
Bromodi ch 1 oromethane
Dichloroacetonitrile
1 , 1 , 2-Trichl oroethy 1 ene
Chlorodibromomethane
1 , 1 ,2-Trichloroethane
1 ,2-Oibromoethane
2-Chloroethyl vinyl ether
2 -Chi oroethy 1 ethyl ether
Bromoform
1,1,2,2-Tetrachloroethane
1 , 1 , 2, 2-Tetrachl oroethy 1 ene
Chlorobenzene
1 ,2-Oibromo-3-chloropropane
MDLA= Method Detection Limit at
MDL a t(n_i
, V" I »
where:
ug/L Percent Percent
0.02
0.025
0.01
0.024
0.0125
0.0125

0.0125
0.0125
0.0125
0.0125
0.0125
0.0125
0.05
0.0063
0.0245
0.020
0.067
0.0502
0.050
0.0515
0.0198
0.0128
0.025
0.058
60
96
60
104
93
103

94
102
100
77
94
94
88
99
94
85
102
106
82
100
84
89
93
84
99% confidence that the
.99) (S)

MDL = the method detection 1
t(n-l,.99)= tne students' t value


imit
appropriate
26
8.5
52
9.2
7.0
4.6

5.5
7.6
7.4
9.3
6.8
5.2
28
3.6
6.8
13
12
39
14
12
20
4.4
7.9
19
value



for a
A. -i 	 -i_ _
MDLA
0.01
0.006
0.1
0.008
0.003
0.002

0.002
0.002
0.002
0.003
0.003
0.002
0.04
0.0007
0.005
0.007
0.03
0.07
0.02
0.02
0.01
0.001
0.001
0.03
is not zero.



MDLB
0.001
0.01
0.03
0.003
0.003
0.003

0.002
0.002
0.002
0.001
0.002
0.003
0.04
0.0006
0.008
0.002
0.04
0.02
0.01
0.05
0.004
0.001
0.005
0.05




99% confidence
. . • -1_1_ — T
                                a
                      degrees of freedom.
       S           =  standard deviation of the replicate analyses
MDLB= Estimated Method Detection limit
                   MDL8 = 3N X RF
       when:
       N  = average noise level at retention time of organohalide
       RF = response factor of organohalide
c = Average background level for methylene chloride 0.1 ug/L

-------
                               References

1.  OSHA Safety and Health Standards, (29CFR1910) Occupational Safety
    and Health Administration, OSHA 2206, (Revised January 1976).

2.  Identification and Analysis of Organic Pollutants in Water,
    Keith,  L. H., Ann Arbor Science, p.  87 (1976).

3.  "Handbook for Analytical Quality Control  in Wastewater and
    Wastewater Laboratories," EPA-600/4-79-019 U.S. Environmental
    Protection Agency, Environmental Monitoring and Support Laboratory
    - Cincinnati, Ohio 45268, March 1979.

-------
              OPTIONAL
              FOAM TRAP
  EXIT 1/4
  IN. O.O.

•— 14MM. O.O.

INLET 1/4
IN. 0.0.
1 4 IN. O.O. EXIT
                                      SAMPLE INLET

                                       2-WAY SYRINGE VALVE
                                       17CM. 20 GAUGE SYRINGE  NEEDLE
                                      6MM.O.D. RUBBER SEPTUM
                                      ^ 10MM, O.D.
                                           INLET
                                           1/4 IN. O.O.
                         1/16 IN. O.O.
                         STAINLESS STEEL
                                                         13X MOLECULAR
                                                         SIEVE PURGE
                                                         GAS RLTER
          10MM. GLASS FRIT
          MEDIUM POROSITY
                                                            PURGE GAS
                                                            FLOW CONTROL
                   FIGURE 1.  PURGING DEVICE

-------
       PACKING PROCEDURE
                                                        CONSTRUCTION
                  MUUIPURPOSE I«AP

        GIASS WOOl  I/MM
ACIIVATED CHARCOAL 7.7CM
GRADE IS
SIUCA GEl   7.7CM
   IfNAX   7.7CM
 351  0V I    (CM
 GIASS WOOl
           SMM
                 \
////////
/////
    7"/fOOI  MSISIANCE
WIRE WRAPPED SOIID _,-—
                               (DOUBLE LAYER)
                                       I5CM
                                     7-/fOOI (IfSISIANCE
                                     WIRE WRAPPED SOUD
                                                                  COMPRESSION NIIINO NU!
                                                                  AND fERRUUS
                                                                       IHERMOCOUm/CONIKOUER
                                                                       SENSOR
                                                                              IUCIRONIC

                                                                              lEMPERAIURf

                                                                              CONIRO1

                                                                              AND     'I

                                                                              PYROME1ER
                                                                   IUBING J5CM 0105 IN. ID
                                                                   OIJ5 IN. O.O. SlAINtESS SIEEl
               IRAP INlif
            FIGURE  2      TRAP

-------
        CARRIER GAS FLOW CONTROL

     PRESSURE REGULATOR
                 LIQUID INJECTION PORTS
  PURGE GAS
  FLOW CONTROL \
13X MOLECULAR
SIEVE FILTER
                    / /
                     /
                    //;
                    /
'/
                                             COLUMN OVEN

                                               CONFIRMATORY COLUMN
                                             TO DETECTOR

                                                 ANALYTICAL COLUMN
                            OPTIONAL 4-PORT COLUMN
                            SELECTION VALVE
                               TRAP INLET (TENTAX END)
                  6-PORT VALVE /   RESISTANCE WIRE
                                                                             HEATER CONTROL
                                                                 NOTE:  ALL LINES BETWEEN
                                                                       TRAP AND  GC
                                                                       SHOULD BE HEATED
                                                     PURCINQ DEVICE     TO 80°C
               FIGURE   3  PURGE-TRAP SYSTEM  (PURGE-SORB MODE)

-------
           CARRIER GAS FLOW CONTROL


   PRESSURE REGULATOR
PURGE GAS     v
FLOW CONTROL  X
13X MOLECULAR
SIEVE FILTER
                                    LIQUID INJECTION PORTS
       wwu4
                                          OPTIONAL 4-PORT COLUMN
                                          SELECTION VALVE
                                                       J
 COLUMN OVEN
f


	CONFIRMATORY COLUMN
TO DETECTOR


"^-^ ANALYTICAL COLUMN
          TRAP INLET (TENAX END)
6-PORT VALVE I RESISTANCE WIRE
      /   ^v   /
                                                      TRAP
                                                      180°C
                                                                      HEATER CONTROL
                                                           NOTE:  ALL LINES BETWEEN
                                                                 TRAP AND  GC
                                                                 SHOULD  BE HEATED
                                                                 TO 80°C
                                               PURGING DEVICE
             FIGURE 4 PURGE-TRAP SYSTEM (DESOR0 MODE)

-------
                                     COIUMN IX If low ON CM»Of»C« •




                                     ttOGIAM  41-C I MINUIH C/MINUII IO 110'C



                                     DIIIC10I HAII ;M A OMIAIINO *| HMO
      •IIINIION IIMI MWIUIIt




CH«OMAIOO«AM Of 0 4u,/| SIANDARO
                                                                                14
                                                                                        It

-------
               ••I- r* c c •
5U;.3y  l'°.- - •"'•' "Ci
                                     VINYl CMIOIIDE * CHLOtOMnMANE
                                  MOMOMETMANE

                            UDfCHtOIOETHTUNE
                                        OilOKOETMANE

                                1-1.3. OICNIOROITNTUNE
                                  MfTKTUNt CWOUOE
                                                        CHtOROrOlM   *  «»-1.2.0tCHlOtOElHYliN€
                                                   SIOMOCHLOKOMHHANE
                                1.I.DKHIOIOETHANE
                                               TIICHIOIOETHANE * l.l.2-I«ICHlO«OnHtlENt
                                            ••OMODICHlOtOMfTHANE
                                                          DtMOMOMEIMANE * Tn«ACHlO*OnMTllNE
                                                                l.2-OICHlO*OElMANt
                                                 DWIOMOCHlO-CHlO»OTOlUtNt
                                                              1.4-aiCMU»O«UTANE * •-MCHIOIOKNZSME
                                                                       HEXACHLOKOMnADONE

                                                       U, «-TIICHLOIOB(NZINE

-------
                                       6MM O.O. HALF-HOLE
                                       CYLINDRICAL SEPTUM
                                      8MM OD.TUBING
                                      9MM LONG
FIGURE 7    MODIFIED VOLUMETRIC FLASK

-------