United States
Environmental Protection
Agency
Environmental Sciences Research
Laboratory
Research Triangle Park NC 27711
EPA-600/8-81-007
April 1981
Research and Development
Operation Manual for
Automatic
Dichotomous
Samplers

Application to
Beckman Dichotomous
Samplers

-------
                                                 EPA 600/8-81-007
                                                 March 1981
OPERATION MANUAL FOR AUTOMATIC DICHOTOMOUS SAMPLERS
    Application to Beckman Dichotomous Samplers
   John D. Spengler, Ph.D., William A. Turner,
     F. Peter Fairchild, and Jane E. Slaughter
          Harvard School of Public Health
    Department of Environmental Health Sciences
            Boston, Massachusetts 02115

                        and

                 Thomas G. Dzubay
    Atmospheric Chemistry and Physics Division
    Environmental Sciences Research Laboratory
        Research Triangle Park, N.C. 27711
              Contract No. D4806NAST
                   Project Officer

                 Thomas G.  Dzubay
    Atmospheric  Chemistry and  Physics  Division
    Environmental  Sciences  Research  Laboratory
         Research Triangle Park,  N.C. 27711
     ENVIRONMENTAL  SCIENCES  RESEARCH  LABORATORY
         OFFICE  OF  RESEARCH  AND  DEVELOPMENT
        U.S.  ENVIRONMENTAL PROTECTION AGENCY
         RESEARCH TRIANGLE PARK,  N.C.  27711

-------
                                   DISCLAIMER

     This  report  has  been reviewed  by  the  Environmental  Sciences  Research
Laboratory,   U.S.    Environmental   Protection   Agency,   and  approved   for
publication.  Approval does not signify that the contents necessarily reflect
the views  and  policies of the U.S. Environmental  Protection Agency,  nor does
mention  of  trade   names  or  commercial  products  constitute  endorsement  or
recommendation for use.
                                        n

-------
                                    ABSTRACT

     The  Beckman  automatic  dichotomous  sampler  is one  of  several  samplers
currently available  that simultaneously collects coarse  and  fine atmospheric
particles.   These  instruments  are  becoming  more widely   used  to  support
research and regulatory  studies  that require information on size fractionated
particles.   The  operating  manual  is  intended  for  use  by  technicians  and
operators of  the  Beckman sampler.   It presents discussions  of  the  nature of
atmospheric  particles and  the  basic  operating  principles  as a  means  of
providing operators with  a  more  complete understanding of  the  instrument and
its uses.

     Factors to  be considered in  site  selection  are  discussed,  followed by
instructions for  installation and  checkout once the  most suitable  site has
been identified.  Special attention is given to checking for leaks in the flow
system,  as  field  experience has  shown this  to  be a potential problem.   Other
areas  of  concern such  as  operating  instructions,  calibration  techniques,
maintenance  schedules,  quality  assurance,  and  trouble  shooting  are  also
addressed.

     The manual should serve  as  an effective supplement  to information given
in  the manufacturer's instruction manual.   Various modifications to original
equipment  and   instructions  that will  simplify  and  improve performance are
described,  including  lists  of  parts  and  suggested   suppliers.   Areas  of
uncertainty about  the design  or  performance are  identified.  The manual will
be revised and updated as further field experience and research  results become
available.

This report was submitted in fulfillment of Contract No. D4806NAST by John D.
Spengler  under  the sponsorship  of the  U.S.  Environmental  Protection Agency,
This report was completed as of December 31, 1979.

-------
                                    CONTENTS

Abstract	iii
Special Note	   iy
Figures	yii
Tables	viii
Acknowledgements 	   ix

     1.   Introduction 	    1

          Nature of Atmospheric Particles	    1
          Description of the Instrument	    3
          Instrument Specifications	    5
          Purpose and Use of Document	    5

     2.   Principles of Operation	    7

          Sampler Inlet	    7
          Virtual Impactor .	    9
          Sampling Flow System 	    9
          Filters	11
          Filter Change Mechanism	14
          Temperature Control System 	   17
          Sampler Control System 	   18

     3.   Considerations in Site Selection	21

          Siting Criteria	21
          Installation Requirements	23

     4.   Installation Procedure 	   25

          Delivery and Unpacking 	   25
          Sampler Installations	25

     5.   Checkout and Calibration	31

          Instructions for Sampler Checkout	31
          Instrument Calibrations	34
          Calibrations Using Rotameters	34
          Calibrations Using A Dry Test Meter	44
          Filter Overload Sensing Switch Setting  	   47

-------
                              CONTENTS  (continued)

     6.    Operating  Procedures 	   50

          Sampler  Housing	50
          Front  Panel  Controls	51
          Interior Flow Controls  	   53
          Operating  Instructions  	   53
          Simplified Operating Instructions	62
          Sampler  Operation	64
          Keypad Operation 	   68

     7.    Quality  Assurance	72

     8.    Maintenance	78

          Schedule for Routine Maintenance 	   78
          Maintenance Instructions - Yearly Maintenance	79
          Maintenance Instructions - Semiannual Maintenance	80

     9.    Troubleshooting	83

References	84

Appendices

     I.    Parts and  Accessories	85

     II.  Correction of Rotameter Readings to Actual
          Pressure and Temperature 	   90
                                        VI

-------
                                    FIGURES

Number                                                                   Page

    1    Size distribution of fine and coarse atmospheric particles.  ...     2
    2    Sampler showing housing and standpipe	      4
    3    Sampler with front cover removed	     8
    4    Sketch of virtual impactor	    10
    5    Impactor collection efficiency versus particle size 	    10
    6    Sampler flow system	    12
    7    Filter tray assembly	    13
    8    Interior of sampler showing filter change mechanism 	    15
    9    Sketch of filter change mechanism 	    16
   10    Temperature control ranges and switch actuating temperatures.  .  .    19
   11    Electronic control system diagram 	    19
   12    Interior of sampler with impactor installed 	    27
   13    Standpipe extension and supporting sleeve 	    28
   14    Fine flow calibration at base using 604 rotameter	    36
   15    Fine flow calibration at inlet using 604 rotameter	    38
   16    Coarse flow calibration at base using 603 rotameter	    39
   17a   Fine flow calibration forms	    41
   17b   Coarse flow calibration forms 	    42
   18    Calibration using Singer dry test meter 	  45
   19    Sampler front panel controls	52
   20    Interior flow controls	54
   21a   Julian calendar for non-leap year 	  57
   21b   Julian calendar for leap years	58
   22    Example of printer printout showing programmed input
         instructions and operational outputs	66
   23    Typical printer output for interrupt cycle	69
   24    Suggested weekly log form	74
   25    Suggested sample flow audit form	75
   26    Virtual impactor dissassembled for cleaning 	  81

   A-l   Viscosity (Centipoise) of air as a function of temperature	  93
   A-2   Percent of flow correction factor for viscosity changes 	  94
                                        Vll

-------
                                    TABLES
Number                                                                       Page
   1     Topographic and meteorological  influences on local  dispersion ...  22
   2     Source configuration and its influence on dichotomous
         sampling	23
                                        vm

-------
                                ACKNOWLEDGEMENTS

     The  authors  wish  to  express their  appreciation  to Beckman  Instruments
Incorporated for  permission to  use  printed materials from  their  instruction
manual.
                                         ix

-------
                                    SECTION 1

                                  INTRODUCTION

NATURE OF ATMOSPHERIC PARTICLES

     This report describes  the  operation,  calibration, and maintenance of the
Beckman  Automatic  Dichotomous  Sampler.   This is  one of  several  dichotomous
samplers available, all of which are designed to collect simultaneously coarse
and  fine atmospheric  particulates.   Both  coarse and  fine particles  are  of
special  interest  because  the  bimodal  distribution  of  particles  in  the
atmosphere  reflects  differences   in  sources  (see  Figure  1).    In  general,
grinding, abrasion,  and mechanical  crushing produce particles  in the coarse
mode  approximately   2.5   urn  in   diameter  and   larger.   Particles   in  the
accumulation of  fine  size  range are in  general  formed from combustion either
directly  or by  gas  to  aerosol  conversion.  These  fine  sized  particles are
believed  to be  of  greater  concern for health effects  because  they will more
readily penetrate to the deeper portions of the lungs.

     Because of  the  differences  in mechanisms of formation,  fine and coarse
particles generally  have  distinctly different chemical compositions.  Many of
the  elements found in the  soil, such as silicon, calcium,  iron, and aluminum,
generally appear in the coarse particle mode. Products of  combustion activity
such as sulfates, nitrates, organics, lead and soot are generally  found  in the
fine particle fraction.

     Once suspended  in the atmosphere particles are  subject to gravitational,
diffusional  and  other  forces  as  well  as  cloud  forming  processes  that
continually modify  their  distribution.   As a result, the  ultra fine particles
are  agglomerating and  growing  in size, thus adding  to  the fine particles in
the  accumulation mode from .1  to  2.5  urn in diameter.  Gravitational settling

-------
C/)
C/)
<
   cc
   HI
   LU
                                     1-0                   10-
                       PARTICLE DIAMETER, jam
  Figure  1.   Size distribution  and  chemical  composition  of  fine and coarse
            atmospheric particles (1)

-------
and  inertia!   impact!on  act  on  larger  particles   to  remove  them from  the
atmosphere.   Although  at times there will  be  particles as large as 50  (jm to
100 urn  suspended in the  lower atmosphere,  the mass of these  coarse fraction
particles is  distributed around  10  urn in  diameter.  The  dichotomous  sampler
was specifically designed to collect suspended particulate  matter separately
in both the fine and coarse size ranges. This is consistent with particle size
distributions  in the atmosphere  and the retention  of  particles  in the upper
and lower human  respiratory tract.

     Since the  dichotomous  sampler collects the fine and  coarse  particles on
separate  filters,   these  filters  can  be   used   in   subsequent  analysis.
Techniques   such  as   x-ray   fluorescence  (XRF)  provide   information   on
approximately  20 elements.   Ion chromatography can  identify sulfate, nitrate,
and ammonium  ions,  among others.   This information  is valuable in  identifying
source  contributions  to  ambient   concentrations  and  the   toxicity  due  to
specific  metals or  compounds.   Sampling of the fine and  coarse  fraction may
aid the assessment of visibility  reduction  problems  and fugitive dusts.

DESCRIPTION OF  THE INSTRUMENT

     This  report  describes  the   operation and  calibration  of  the  Beckman
Automated  Dichotomous  Particulate Sampling System, hereafter  referred to as
the dichotomous sampler  or the  Beckman  sampler.  All components except for the
inlet  are housed in  a weather  resistant,  internally  heated enclosure, shown in
Figure 2.  Air  is drawn into  the instrument through the  fractionating  inlet,
which   is  designed  so  that  particles  larger than  approximately  15  urn are
excluded.   The air  sample  next flows  through  a  virtual  impactor,  which
simultaneously  separates  the particles  into  a   fine  fraction  and   coarse
fraction.   The  two  size fractions are  then deposited on  separate  filters for
subsequent  analysis.  Air flow through the sampler is maintained  by separate
fine and  coarse flow systems.

     An  important  feature   of   the  Beckman  dichotomous  sampler   is   its
microprocessor control  system, which allows for automated  continuous sampling.
Two filter  trays, each containing 36 filters,  are  placed  in  the sample  changer

-------
                                                            FRACTIONATING INLET
                                                           SAMPLING STANDPIPE
                                                          INSTRUMENT SHELTER
Figure 2.  Beckman Automated Dichotomous  Sampler,  Showing Housing,  Standpipe,
          and Fractionating Inlet in  Operating  Position

-------
 (see  Figure 7).   The sampler can  then  be programmed to collect automatically
 35  pairs  of samples, at preselected times and sampling  intervals.  (One pair of
 filters  is  always  maintained  unexposed  as a  control.)   A  heater/thermostat
 system maintains  the internal temperature at 33°C and permits  sampling through
 variations  in  ambient temperatures of -40°C to  +50°C.

      A  more  detailed  discussion  of  instrument   subsystems and   operating
 principles  is  provided  in  Section  2 of  this report.  Instrument specifications
 are given below.

 PURPOSE AND USE OF  DOCUMENT

      This report  is primarily  for use  by those responsible  for  establishing
 and operating  dichotomous  sampler  networks  and  for  field personnel  responsible
 for maintenance and calibration of the  Beckman  sampler. The report is designed
 to  provide  background  information and descriptions of how the sampler  works
 and suggestions  for its use.   Considerations  for site selection  are provided
 along with  installation  instructions.  Calibration, operation and  maintenance
 procedures   are  presented,  along  with suggested quality  assurance and  data
 handling  procedures.   The information is  based on materials provided by  the
 manufacturer along with the  results  of extensive field use of the instrument.
 It   is  expected  that  the manufacturer's  instruction manual  (2),  including
 trouble   shooting  procedures  and instrument   schematics  diagrams,  will   be
 available to the  user  of  this report.

 SPECIFICATIONS

Inlet Upper  Cutoff Point (50%):         15   urn  nominal;  12  to 18 urn  for wind
                                       speeds  below 20  km/hr
Virtual  Impactor  Cut Point (50%):      2.5  urn
Virtual  Impactor  Losses:                <5% for particles between 1  and 15   pm
                                       aerodynamic diameter; <15% at cut point
Sample Flow Rate:                       16.7 £/min total  (1 cubic meter per hr)

-------
Sample Flow Stability:
Timing Accuracy:
Start Time Selction:

Delay Period Between Samples
or Sample Groups:
Number of Filters per Group:
Filter Material:
Filter Size:

Number of Filters:
Housing:
Ambient Operating Temperature Range:
Operating Humidity Range:
Power Required:
 Power Failure:

 Power Cord:
 Dimensions:

 Instrument Weight:
 Shipping Weight:
±5% per week
±2 min per day
Presettable to any number of days,
hours, or minutes up to 365, 23, and
59, respectively
Presettable to any number of days,
hours or minutes up to 364, 23, 59
Presettable from 1 to 36 filters
I  urn Teflon standard, 2 urn optional
29 urn deposit dia. on 5 cm x 5 cm
square frame
36 pair in 2 trays
Weathertight; internally temperature
controlled at 33°C
-40°  to +122°F  (-40° to +50°C)
0  to  100% RH
115 VAC, 60 Hz;  200 VA nominal  at  25°C
with  pumps off;  300 VA max  at  25°C with
pumps on; 850 VA max with  heaters  and
pumps on
Battery pack maintains timekeeping
function  for 30 min should  power  fail;
time  of power  failure  is  printed
8 ft  (2.4 m)  long,  U-ground plug
40 in.  H  x  26  in.  W x  19% in.  D (102 x
66 x  50 cm);  plus inlet  protrudes 20
 in.  (51  cm)  from top
 175 Ib (80  kg)
 220 Ib (100 kg)

-------
                                    SECTION 2

                             PRINCIPLES OF OPERATION

     This section gives an  explanation of the principles of operation for the
six main components of  the  dichotomous sampler:   fractionating inlet, virtual
impactor, sample  flow system,  filter  system, temperature  control  system and
sampler  control  system.   Figure  3  shows the major subassemblies which are
readily visible with the front cover removed.

SAMPLER INLET

     The air sample enters  the fractionating inlet, which  protrudes  from the
top of  the  sampler housing as shown  in Figure  2.   Because the  bell  shaped
inlet  has  axial  symmetry,  the  sampling  efficiency is  independent  of  wind
direction.   With  air  entering  through the  bottom of the inlet,  rain is  kept
out,  and  a coarse  screen  keeps  large  particles  and insects  from  being
entrained.  The inlet is designed so that particles greater than 30 urn will not
enter.   Particles  between  about  15  pm  and 30  (jm have, a  settling  velocity
roughly  equivalent  to  the average upward velocity  of  the inlet so that these
particles settle out  of the  inlet and do not enter the impactor (3);  thus the
inlet 50% cutpoint  is 15 pm,  i.e., the  inlet is approximately 50 % efficient
at collecting 15 pm particles.  It should be noted that wind speed does affect
the inlet performance.  Efforts are underway to study the effect of wind speed
on  size  cutoff, and  operators should check  latest  information  on wind speed
dependence and inlet design.

-------
     SAMPLER CONTROLS
                                                                 VIRTUAL IMPACTOR
                                                                     FILTER TRAY
                          FLOW SYSTEM
Figure 3.  Beckman Dichotomous Sampler with Front Cover Removed Showing Major
           Subassemblies

-------
VIRTUAL IMPACTOR

     The  sampled  particles  smaller than  15  (jm pass  through an  inlet  tube,
which serves to straighten the flow, and then into the virtual impactor,  which
is shown  in  Figure  4.   The virtual impactor is the basis for the operation of
the dichotomous sampler  and  was designed by Loo et al(4).   Particles entering
the  impactor are  accelerated  through  a  nozzle,  and enter  the  fractionation
zone.   Because of  their greater  inertia,  particles  larger  than 2.5 pm are
drawn  directly into the  collection nozzle and then  deposited onto  a filter.
The  large particles do  not  follow the flow stream and  "impact"  into the low
flow  region   of  the  collection  tube.   Because  of  the  geometry  of  the
acceleration  nozzle,  only  about   10% of  the  total flow  which enters the
impactor  goes  through  the collection  nozzle.  The  remaining  90% of the  flow,
which contains the smaller particles (fine fraction), passes around the nozzle
and through  a  tube to a  separate filter.

     The  cut point  between fine and coarse  fractions depends on flow rates  ,
the  diameters of the  acceleration and  collection  nozzles,  and  the distance
between   nozzles.    Figure  5  shows collection  efficiencies  for the  coarse
particle  fraction.   The  lower 50%  cutpoint  for  the coarse fraction occurs at
2.5  urn.   Losses  of particles  on the walls of the  impactor are measured  to be
less  than 10% at the cut  point, which represents very  good  performance  for  a
single  stage virtual impactor  (4).

SAMPLING  FLOW SYSTEM

      Atmospheric  particulates are  sampled through the  inlet at at  rate of  1
cubic  meter per hour,  or  16.7 liters  per minute. The sample  flow entering the
impactor  is split into  two  separate flow systems (see Figure 6). About 10% of
the  sample goes through the  collection  nozzle and  onto  the coarse filter at  a
rate  of  1.7 liters  per  minute.   From the  filter,  the  air sample is carried
into  a  ballast volume  which  smooths the  flow  for the  WISA vibrator  pump.
Control  and flow indication  in  the coarse flow  system is  provided by a needle
valve and rotameter.

-------
                                        AIR FLOW
                  SMALL PARTICLE
                     TRAJECTORY


                  LARGE PARTICLE
                     TRAJECTORY
                                                        ACCELERATION NOZZLE
                                                        FRACTIONATION ZONE

                                                        COLLECTION NOZZLE
                                     TO COARSE
                                     FRACTION FILTER
TO FINE
FRACTION FILTER
Figure 4.   Sketch of Virtual  Impactor Showing  Separation of  Coarse and
             Fine Flows
              100
               80
              •60
               40
               20
                                      COLLECTION EFFICIENCY
                          23       5    7    10

                             AERODYNAMIC PARTICLE DIAMETER, i
        20
              30
Figure  5.   Graphic    Representation of  Virtual  Impactor Collection
             Efficiency  vs. Particle Size for  a Design  Cutoff of  2.5 ym
                                        10

-------
     The smaller sized finer particle flow passes around the collection nozzle
and is collected  on  a second filter at a flow rate of 15.0 liters per minute.
After  the  filter,  air  goes through  a  flow control  device which  provides
regulated flow in conjunction with a precision needle valve.  The flow setting
of 15  £pm  is  initially  set by  the needle  valve  and  rotameter.   The needle
valve  then  functions as a  fixed flow control orifice.  A  dual  diaphram pump
provides the vacuum necessary to induce flow through the system.

     A differential pressure switch acts as a safety release in case of filter
overload.   The  switch   is  connected  across  the  needle valve  to sense  the
loading  on  the fine  filter (indicated as "PS3" in  Figure  6).   If resistance
across the  fine  filter  rises enough to cause a 5% decrease in regulated flow,
then the next pair of filters is automatically inserted.  This is likely if an
unusually high  atmospheric  particulate loading exists, such as in an episode.
The  flow controller  maintains  a  constant  flow  rate  until  the  fine filter
loading  causes a  pressure  drop of about  20 cm  (8 in) Hg.   Experience has
indicated that the filters  are effectively fully loaded at  that point (2).  It
should   be  noted  that  earlier  production  models  were  equipped with  two
additional  pressure  switches  to  detect torn  or  missing  filters  ("PS1" and
"PS2"  in  Figure  6).    Later  models  of the  instruments  do  not  have these
switches.

     Accurate  flow control  is extremely  important  in the dichotomous sampler,
as  the  pressure  drop  across  Teflon filters  can  increase  rapidly  as they
collect  particles.   Maintaining  the 16.7  £pm  flow  through  the  system  is
essential to maintain the particle  size  cutoff specifications of the inlet and
virtual  impactor.

FILTERS

     The dichotomous  sampler uses Teflon  filters mounted  in 5 cm x  5 cm frames
with  a 29  urn  deposit dia., shown in  Figure  7.  Filters are  available in two
types,  one  in  which the filters  are bonded  to the frames,  and a second in
which  the filter is removable to  facilitate subsequent analyses.  The  filters
                                      11

-------
      PRESSURE
       SWITCH
                      FRACTIONATING
                      INLET
                                                   THOMAS DUAL
                                                DIAPHRAGM PUMP
  DIFFERENTIAL
PRESSURE SWITCH
                                                                                         TOTAL
                                                                                         SYSTEM
                                                                                         FLOW - 16.7
                                                                                         LITERS/M1N
                                                                                             RECOMMENDED
                                                                                             INSTALLATION
                                                                                             OF VACUUM GAUGE
                                       HISA PUMP
                                                          NEEDLE
                                                          VALVE
Figure 6.   Sketch of  Dichotomous Sampler Flow System Showing Fine  and Coarse Flow Paths.

-------
FILTER TRAY
                                                             FILTER TRAY PREPARED
                                                             FOR SHIPMENT
                                      FILTER HOLDER
                FILTER INDEXING
                GUIDE
Figure 7.  Filter Tray Showing Filter Holders and Indexing Guide
                                       13

-------
have a 1  urn  pore size,  and are preferred  for their low initial  pressure drop
and minimal water retention.   Also the Teflon material  has negligible particle
penetration and  efficiently collects  particles as small as  0.04  pm.   The low
                                2
weight  per  unit area  (1 mg/cm  )  and  chemical  inertness  make  the  filters
suitable for XRF, p-ray densitometry and wet chemical analysis.
     The filter frames are held in trays having 36 slots each.   Two trays, one
for the  fine  fraction and one for the  coarse  fraction, are inserted into the
sample change mechanism,  as  shown in Figure 8.  The sampler can be programmed
to collect automatically up to 36 pairs of filters, but the first pair must be
held blank to serve as a control.  Pairs of filters are advanced by the sample
changer  when  the  preselected delay period between  samples  has been completed
or  when  the  flow  differential  pressure switch has  been  activated because of
heavy loading on one of the filters.

     Each filter is weighed and  labeled with an identification number prior to
shipment of  the filter trays to  the sampling  site.  It is extremely important
that  complete  records  be  maintained  of  the  samples  used  and  of  sampler
operation.  A   suggested   data   collection  and  record  keeping procedure is
provided in  Section 7; however,  the operator  should make sure that the  forms
and instructions  appropriate to  the particular  sampling  program  are on  hand.

FILTER CHANGE MECHANISM

     The sample changer subsystem consists of a filter shuttle drive,  a tray
 indexer  and   a filter  seal mechanism,  as shown in  Figure  9.   These  three
mechanisms   are mechanically  independent,  but   all  are  controlled  by  the
 instrument  microprocessor.   The  filter  shuttle  moves  back and forth,  pushing
 the clean filters  into position to  collect  a  sample and returning the used
 filters   to   the   tray.    This   shuttle  is  attached   to  a  simple  eccentric
 cam-actuated rotating  assembly  which moves the  shuttle  in and out of  filter
 trays.   Feedback for  the  position of  the rotating assembly is provided by two
 mechanical  limit  switches actuated by  a pin  on the assembly.
                                      14

-------
         CARRIAGE
         RELEASE
         KNOB
                                                                      FILTER
                                                                      SHUTTLE
FRONT OF INSTRUMENT
FINE PARTICLE
FILTER
COARSE PARTICLE
FILTER
  Figure 8.  Interior of Sampler Showing Filter Change Mechanism, Filter
             Trays Inserted and Filter Pair in Place (Virtual Impactor has
             been removed)
                                      15

-------
                                                             FILTER  TRAYS
      RECIPROCATING
            SHUTTLE
SHUTTLE DRIVE
    ECCENTRIC
SHUTTLE DRIVE
        MOTOR
                                                                   CARRIAGE RELEASE KNOB


                                                                   OPTICAL DETENT
FILTER SEAL
ECCENTRIC
                          JACKSCREW
       FILTER SEAL
       MOTOR
                                                                                 MOVABLE CLAMPING
                                                                                 PLATE
                                                                             OPTICAL DETENT
                                                                             PICKUP HEAD
                                                                   TRAY DRIVE MOTOR
                           Figure 9.  Sketch of Filter Change Mechanism

-------
     The  filter  shuttle  drive  only  operates  in  the  horizontal  direction.
Vertical movement of the filter trays is controlled by the tray index assembly
so that  fresh  pairs  of filters can be loaded in proper sequence by the filter
shuttle mechanism.   Vertical  motion  is provided by a jack-screw which engages
the tray  index  carriage.   A knob at  the  top of the carriage can be turned to
disengage the  carriage  from the jackscrew to allow initial positioning of the
trays.  Unlike  the  other  two mechanisms in the sample changer, the tray index
assembly  has multiple  rest positions.   Position  feedback for  this assembly
therefore  is provided  by  a multiposition  optical  limit switch  rather than
simple  mechanical  limit switches.  In Figure 9  the notched optical detent and
pickup  head are  shown on   the  front of  the  trays  for  clarity;  however,the
actual  location is behind the tray, since access is not normally required (2).

      The  filter seal mechanism  clamps and  seals  the  filter  slides into the
pneumatic  system  once the  shuttle  drive  has  pushed  them  into  place  for
sampling.    It  is   operated in  sequence  with  the  other  mechanisms  under
microprocessor  control.  Vertical  motion is  applied to  the  filter clamping
tubes by a simple eccentric  cam.  Position  feedback  is provided through limit
switches  actuated by two eccentrics on the end  of  the  clamp shaft  (2).

      Figure  8  shows  the filter change mechanism with  filters in place.

TEMPERATURE  CONTROL  SYSTEM

      Sampler  housing   is   internally   heated  to   +33°C   (91°F)   using   a
fixed-set-point,   250-watt,   proportional   temperature   controller.    This
controller   will  maintain  the  housing  at  the   control  temperature  through
variations   in  ambient  air  temperature  from  +30°C  to  -40°C.    If  ambient
temperatures rise  above  the  controller  set   point  (+33°C),  the temperature
inside the  housing will rise accordingly.

      Besides the  proportional  temperature  controller,  several  other  thermal
switches are included  in the  system  to help control  the internal temperature
and  to protect circuitry  and components:
                                      17

-------
          Where  ambient  temperatures  are  very high,  a  high-temperature  thermal
          switch will  actuate at +48°C  (+118°F)  to  energize  the  intake blower
          fan.  The fan directs air flow  over the microprocessor  circuit board
          and associated electronics  to  prevent extended operation at highly
          elevated temperatures.   This thermal  switch  is  located inside  the
          electronics  housing assembly  on the  filter changing  assembly  (2).

          A low-temperature thermal  switch disconnects  both  the  battery power
          and the +5 Vdc regulated power to  the microprocessor  when  internal
          housing temperatures go below  0°C.   The instrument must warm  up  to
          +4°C (39°F)  before power is  restored.  This thermal  switch is  located
          within  the  electronics housing on  the  interconnect  circuit  board
          (2).

          A third thermal  switch, mounted on the underside  of the instrument
          base plate,  closes at -7°C.   This connects a second 250-watt  heating
          element  to  the  controller,   to   increase  internal   heating  for
          operation in extremely  low  ambient  temperatures (2).

     Figure 10 shows  the thermal  switch actuating temperatures and the  control
ranges for the  Sampler.  Note that a neon indicator lamp,  which  is visible on
the front side of the power supply,  is connected across the heater elements to
indicate when the heaters are fully on (lamp  on continuously), in proportional
control  (lamp flashing), and off  (lamp off).

SAMPLER CONTROL SYSTEM

     One of the most unique features  of the Beckman dichotomous sampler is its
Microprocessor  Control   System,   shown  in  Figure  11.   The  microprocessor
generates  a  sequence   of  timed  comands  which  control  the visual  display,
lights,  keyboard, printer,  motors, and pumps.  The total sampler operation is
controlled  by  an 8-bit  microprocessor-based  system  using  an   8085  Central
Processing Unit (CPU).   Control  programs are  stored in 4K of Read-Only memory.
                                     18

-------
                   TEMPERATURE
                   CONTROLLER
                    INTAKE FAN
                   MAX HEATER
                       POWER
                      BATTERY
                         AND
                     tP POWER
                                                  33°C
                                  FULL ON
                                               PROP
                                        OFF
                                          250 W
                               500W
                                   -7-C
                                               QN
OFF
    -7*C +4°C (Ta)
                          -40°C
                                   Tr INTERNAL TEMPERATURE

                                   T - AMBIENT TEMPERATURE
                                    A
                                                         OFF
                                                         ON
                                                           +48"C
Figure 10.   Diagram  Showing Temperature Control  Ranges and  Switch-
              Actuating Temperatures
KEYPAD
DDDD
DDDD
DDDD
DDDD



POWER
SUPPLY

BATTERY
BACK-UP
DISPLAY
LIGHTS



1

— o o —
— o o —
— 00 	
— o o —
— o o —
— o o 	
	 o o 	
f
MICRO-
PROCESSOR







|

SAMPLE CHANGER
FILTER
SHUTTLE



FILTER
SEAL
t
VERTICAL
TRAY
t
MANUAL SWITCHES

^
PRINTER



REAL TIME
CLOCK

PRESSURE
SWITCH




PUMPS



Figure 11.   Simplified Electronic Control  System Diagram

                                       19

-------
The microprocessor  board power  requires  a single +5Vdc  supply,  with battery
back-up (2).

     Sampler operation  is programmed  by the key pad.   Current  data and time,
sample  start  time and  date,  number  of samples, and time  between samples are
entered,  following  the  sequence  indicated by the display  panel  lights.   The
printer provides  a  record  of sample parameters.  Manual switches are included
to actuate the filter shuttle, filter seal, vertical  tray index mechanisms and
the sampler pumps.

     A  more  detailed  description of  the key  pad  display lights  and manual
switch  functions  is included  in Section 6, Operating Procedures.
                                      20

-------
                                    SECTION 3

                        CONSIDERATIONS IN SITE SELECTION

SITING CRITERIA

     Selection of the  sampling  site and sampler location  is  dependent on the
intended use of  the  collected data, as well  as  several  other considerations.
This section provides criteria and a suggested methodology for site selection.
The  information  is  based on the  recommendations  of  the EPA's  Standing Air
Monitoring Work Group (SAMWG) and the final regulations on ambient air quality
monitoring (5). Appendices D and E of reference (4) discuss network design and
sampling probe siting  criteria,  respectively, and  should be reviewed prior to
establishing the monitoring site.  Another valuable reference is the report of
Ludwig, Kealoha and Shelar (6).

     A  dichotomous  sampling program  can serve a  variety of objectives.   The
dichotomous  sampler  gives  information  on  the  coarse and  fine  fraction  of
suspended atmospheric  aerosols  below 15 urn  in diameter.   These fractions may
have  similar or different  sources.   These  sources  may  be  local or distant.
One  size fraction  may  reflect  variations  in local  sources  and dispersion
conditions while the  other fraction does not.  Therefore,  in choosing a site
consideration  must  be  given  to  factors  that  might bias  measurements and
compromise sampling objectives.

     In  siting dichotomous samplers two general categories  of impacts should
be  kept  in  mind.  Meteorological (mesoscale  and microscale)  effects and  local
source configurations might influence fine and coarse  particle  levels.
                                     21

-------
      Meteorological   factors  on  the  local  scale are  the  altered  wind  flow

 patterns or  mixing  conditions by  topographic features.   Tables  1 and  2 are

 included to highlight these considerations.


TABLE 1.   Topographic and Meteorological Influences on Local Dispersion
Topographic Features
Effects
1.  Elevated regions
2.  Deep valleys
3.  Undulating regions
 4.   Regions  of  tree  cover
 5.   Bodies of water
a.   Increased wind speed (and increased
    ventilation) over hill tops.
b.   Occasional impacting of elevated plumes
    on ground level.

a.   Channeling of wind flow along the valley
    axis, resulting in higher average
    concentrations in the valley.
b.   Development of stable, drainage winds
    during calm, nighttime conditions,
    resulting in higher concentrations along
    the valley floor.

a.   Increased atmospheric turbulence near the
    ground level during time of moderate or
    strong winds.  This results in lower
    pollutant concentrations at locations
    near sources.
b.   Accumulation of pollutants in low spots
    during calm, nighttime conditions (i.e.,
    localized drainage wind conditions).

a.   Enhanced  turbulence near the ground
    during moderate or strong winds,
    resulting in lower concentrations for
    locations near sources.
b.  In  fully  covered  regions, blockage of
    elevated  plumes,  resulting  in lower
    concentrations at ground  level.

a.  Increased moisture content  in the  local
    atmosphere,  favoring  fog  formation at
    low-lying spots,  and  affecting  the
    removal  rate of  SOp  and  other pollutants
    from the  atmosphere.
 b.  For larger  bodies of water,  formation  of
     local  circulation (lake  and sea breezes)
    which can cause  ground level  fumigation
    on the landward  size of  sources, during
     sunny daytime  conditions.
                                       22

-------
 TABLE 2.   Source  Configuration  and  its   Influence  on  Dichotomous  Sampling
Source
Effects
1.   Unpaved roads, sanded or salted
    open fields, stockpiles
    sand, etc.

2.   Heavy traffic on nearby road
3.  Overhanging trees -- beneath
    densely forested canopy

4.  Rooftops, gravel roofs
5.  Building wakes
Resuspension of surface dust roads,
contributing to coarse fraction.
Resuspension of coarse particles, local
source of fine particles high in Pb,
Br.

Some possible filtration of coarse
particles; also production of pollens.

Building exhaust vents, local chimney
emissions. Possible resuspension of
roof material.

Possible influence of wake capture of
building emissions.
 INSTALLATION REQUIREMENTS FOR THE DICHOTOMOUS SAMPLER


      Installation  requirements  for  the dichotomous sampler are minimal. It is
 designed  to  operate outdoors, through a range of ambient temperatures of -40°
 to  122°F.   The internal heating blower system essentially eliminates the need
 for  external  temperature  or  humidity control.   Although the  instrument is

 housed  in a weather-tight shelter,  it may  also be placed indoors.  This will
 facilitate   calibration  and  maintenance  and  protect  the  filters  during
 handling.   Indoor  installations  require the  availability of a  level, flat,
 rigid  surface  to support the four  legs of  the  shelter, and some modifications
 to  the sampler inlet standpipe.  Indoor  installations  also  require access to

 the  ambient air  directly  above the sampler to  allow placement  of a straight
 pipe to the fractionating inlet.   Provisions  should  be made to provide enough
 clearance above  the  sampler to  allow removal  of  the shelter which  must be
 lifted  up and  over the  instrument.
                                       23

-------
     The  instrument  weighs  less  than  200 pounds  and can  be  lifted by  two
persons using the  grip  handles  on the sides and rear and the upper lip of the
front  door  of  the shelter.   It can  be  plugged into  any grounded  115  Vac
electrical outlet and requires no further utility connections.

     As with  the placement  of  other particulate samplers, the  inlet  for the
dichotomous sampler must  be  located 2 to 15 meters above the ground (5).   Two
meters is considered  minimum height necessary to avoid entrainment from dusty
surfaces; the  15 meter  upper limit  is  essentially  a  compromise  between the
desire  to  have  instruments  that  are  most  representative  of  population
exposures and the practical considerations of site security and accessibility.
Further,  the  inlet  should  not  be  within  two  meters  of  walls  or  other
obstructions or  within  20 meters  of large  trees  which often provide surfaces
for particle deposition and restrict airflow.

     Since the  dichotomous  sampler  may be  co-located  with hi-vol  samplers it
is recommended  that  the dichotomous inlet  be placed  so that it is not within
two meters of  and at least  0.5 meters  above  the hi-vol. This will reduce the
collection of  copper  and carbon particles  from the hi-vol motor. If possible,
exhaust from the hi-vols  should be ducted away from the dichotomous sampler by
using  a large flexible  hose  or  similar device attached  to  the  bottom of the
hi-vol motor housing.
                                      24

-------
                                    SECTION  4

                             INSTALLATION PROCEDURES

DELIVERY AND UNPACKING

     Although  the  sampler  is  inspected  and  packed to  protect  it  during
shipment, there  is always  the  possibility of damage.  Carefully  inspect the
shipping package  for any external  signs of damage,  note  this  information on
the  delivery receipt, and  have the  delivery  person sign  all  copies  of the
receipt before leaving.   Unpack the instrument and check each item against the
packing slip to make sure everything has been included in the shipment.

     As  soon  as  possible,  the sampler should be  assembled and checked out to
verify that there  is no internal damage.  Experience has shown that the filter
change  mechanism   is  more  likely  to  be  damaged  in  transit  than  other
assemblies.   Also, the shelter  housing legs may be bent  or  broken,  and the
plastic  caps  used to plug the former location of two pressure switches in the
flow  lines  (see  Figure 20) may become  dislodged  or lost.   These caps must be
in place before sampling.

      If  there is  any indication of either  internal or external loss or damage
to the  sampler,  the shippping company and  manufacturer's  sales representative
should be contacted immediately before any  further  use of  the sampler.

SAMPLER  INSTALLATION

      Before  starting the actual installation, make sure the site  location and
instrument  placement  meet the criteria  and  requirements described  in Section  3
of this  report.
                                      25

-------
Outdoor Installation

     The  Beckman   dichotomous   sampler  requires   no   special   installation
procedures if it is to be used outdoors.   The instrument should be placed on a
firm,  level  supporting surface  and  securely anchored.   High  wind  speeds  in
excess  of 50  mph  can  cause the  sampler  to  tip over.   Be  sure  to  provide
several  inches  of  clearance  below the sampler  housing, as the  interior air
circulation ports  are  loacted in the base  plate.  The  fractionating inlet is
attached  to  a sampling  standpipe,  which is  inserted through the  top  of the
sampler  housing.   The  standpipe  should be  lined  up  and firmly butted against
the  impactor  inlet.  Use  a flexible  sleeve  or  a 2-inch length of IV I.D.
tygon  tubing  as  shown in Figure 12 to connect the standpipe and the impactor.
The  clamp located on the underside of the  shelter housing top should then be
tightened to  hold  the standpipe  in place during sampling.  There should not be
any  space between  the standpipe  and impactor in order to maintain a  smooth air
flow into the  impactor.   Once  the  standpipe  and  inlet are  installed, the
sampler  may  be  plugged into  a grounded  115 VAC electrical outlet,  it is then
ready  for checkout and operation.

Indoor Installations

      Although the dichotomous sampler is designed for  use outdoors, it  can be
 located   inside  a  building,   instrument   shelter   or   trailer.    This  type
 installation   requires   minor modifications  of  the  instrument  but   may  be
preferable   in   some   sampling   situations.   If   an  indoor   installation  is
 selected, the fractionating  inlet must  be  located in the ambient air directly
 above the impactor.  This  requires  an extension to  the  inlet  standpipe and  a
 fixed supporting  sleeve  through which  to  raise and  lower the standpipe.   A
 typical  installation is  shown in Figure  13.

      The standpipe extension should  be  made of rigid piping with a nominal  1"
 inside diameter and Ik"  to 1 3/8"  nominal  outside diameter.  The  standpipe  and
 extension together should  be long enough so that the sampler  inlet will be at
 least 2  meters  above  the sampling station roof when the unit  is  in operation.
                                      26

-------
TRAY INDEXING
  TAPE
                                                              STANDPIPE  CONNECTOR
                                                              (TYGON TUBING")
                                                           IMPACTOR
                                                           ASSEMBLY
           IMPACTOR ASSEMBLY
           SUPPORT BLOCKS
    Figure 12.
Interior of Sampler Showing Filter Tray and Virtual Impactor
Installed.  (Note section of Tygon tubing at top of Impactor
for connecting to standpipe; also note tray indexing on side
of filter tray.)
                                         27

-------
                                                    FRACTIONATING INLET
                                                        STANDPIPE
                                                   STANDPIPE EXTENSION
                                                     SUPPORTING  SLEEVE
                                                    SAMPLING STATION ROOF
Figure 13.   Standpipe Extension and Supporting Sleeve as
            Required for Indoor Sampler Installations
                           28

-------
The extension may be attached to the standpipe by using a section of IV tygon
tubing or other suitable material, clamped to the standpipe and the extension.
Be sure that the extension and standpipe are firmly butted together so that no
flow irregularities are induced.

     Material  selected  for   the  extension  tubing  is  a  concern  because  of
potential  wall  losses  along the additional  length.   Stainless  steel  tubing
will reduce this potential, but may not be practical or readily available.  If
PVC  pipe  is  used,  large  wall  losses  can  be  expected.   Aluminum pipe  is
satisfactory if great  care is taken to prevent  aluminum (from the pipe walls
and especially pipe ends) from falling into the sampling stream.

     Since  the  standpipe must  be  free  to  be  raised  and  lowered  to the
impactor,  it cannot  be fixed to  the roof  of the sampling station or trailer.
A  rigid  sleeve which  is firmly  attached  to  the  roof  must be provided (see
Figure 13).  Make  the  sleeve out of PVC piping to minimize the possibility of
metal  particles becoming  dislodged as  the  standpipe  is moved.   The  inside
diameter  should be  sufficient to  allow  free movement  of the  standpipe and
extension  once  they  are joined together.   The supporting pipe should be tall
enough so that the inlet  rests  on  the top of the  pipe  during sampling.   The
pipe may  be fixed to the roof by  metal  brackets such as in Figure 13 or by guy
wires.

     Once  the   standpipe  extension  and supporting sleeve are  installed, the
standpipe  should be lowered onto  the impactor and connected by  use of flexible
sleeve or section  of  Ik"  I.D.  tygon tubing, as  was shown  in  Figure 12.   The
clamp  located  on  the underside of the sampler housing top should be tightened
to  hold  the standpipe in place  during  sampling.   Be  sure that  there  is no
space  between  the  standpipe and  impactor in order to maintain  smooth air flow
into the  impactor.   Once the  standpipe is attached, the sampler can be plugged
into a grounded 115 VAC electrical outlet  and is  then ready for checkout and
operation.
                                     29

-------
     For the  convenience of the operator,  it is suggested that a  muffler  be
installed on the  pump  exhaust  to reduce noise level  in the sampling building.
Several  types  of mufflers  are  available.   Appendix I describes one  of these
that has proven to be effective in  field tests.

     Note:   It is  strongly  recommended that a worm  drive  circulator  clamp  be
used to  hold  the  standpipe  and impactor inlet together. The  clamp should  be
placed around  the two  inch tygon  tubing  section which  is  used to  join the
inlet and impactor.

     Description  and  sources for accessories and materials discussed in this
report are provided in Appendix I.

                                     CAUTION

Always  turn  the  POWER  switch  off  before  disconnecting  the  power  cord.
Otherwise, the battery  backup  system will be activated; after 30 minutes, the
batteries will be completed discharged.
                                      30

-------
                                    SECTION 5

                            CHECKOUT AND CALIBRATION

     Since performance of the virtual impactor is highly dependent on the flow
rate through  the  sampler,  it  is critical  that precise flow  calibration  and
accurate flow control  be  accomplished.   Thus, it is essential  that the system
be  carefully  checked  out and  calibrated  prior to  initial  sampling  and  at
regular intervals thereafter as  suggested in Section 8  of  this  report.  This
section provides instructions and  recommendations  for initial  checkout of the
sampler.  It  also  includes  procedures  for calibration of the  fine and coarse
flow rates,  and  for  total  flow  system checks.   To perform  the  tests  and
calibrations described here, the operator must be familiar  with the operating
procedures described in Section 6.

INSTRUCTIONS FOR SAMPLER CHECKOUT

1.   Once the  instrument is  unpacked  and  set  up,  an operation  check of the
     printer and system  controls should be performed to verify that no damage
     has  occured  to  the internal  mechanisms.   The  operator  should complete
     Items 2  through  11  of  the operating instructions in Section 6, and check
     carefully  that  the light  sequence,  printer  output, clock  and other
     controls are  working properly.   It is not necessary  to  unpack and load
     the filter trays  for this checkout  step.

2.   The virtual impactor assembly should be removed  and cleaned in accordance
     with the instructions in Section 8.  Once the  impactor is reassembled and
     installed, the  impactor inlet should be covered with  a  #7*5 or #8 rubber
     stopper, aluminum foil, or a similar device so that dirt and other
                                     31

-------
     foreign objects will  not inadvertently  enter the impactor.   NOTE:   The
     inlet  should  always  be covered  when the sampling standpipe  is  removed.

3.    It is  strongly  recommended  that  a vacuum type pressure  gauge (see parts
     list in Appendix  I)  be installed in the  flow sytem,  located as  shown in
     Figure 20.   To install, locate the tee type fitting which is at the lower
     right  of  the  fine flow  rotameter.   Note  that  one part  of the  tee  is
     plugged.   Remove  the  plug and install the vacuum gauge.   Be sure that no
     "stringy"  pieces  of  Teflon  tape are left  in the tee fitting after the
     plug  has  been removed.  Position the gauge dial  so that  it  can be read
     easily at the same time as the rotameters.

4.    Since  the flow  checks and calibrations described below will be performed
     with a pair of filters (not the first set, #1) in place,  each filter tray
     should  be  checked  for damaged  filters.   Early models  of  the Beckman
     sampler  contained pressure  switches to  sense torn or  missing  filters.
     However, later production models do  not have  this feature and each filter
     must be checked prior  to loading of  the trays.  NOTE:  The filters should
     be  handled   carefully using  the filter  frames.   The Teflon collecting
     surface should  never  be handled  directly.

5.   To  assist  in indexing the  filters  in the  trays,  it  is  suggested that  a
     piece  of  tape  be placed along  the front  edge  of the  filter  tray, and
     markings  located  on  the tape to  correspond with the filter  numbers  1-36.
     Figure 12 shows a typical installation.

6.   All  flow checks  and  calibrations  should be  performed after  the  sampler
     door has been  closed and top unit  turned  on for  at least 2  hours.  This
     will  allow  the  instrument to reach  its  normal operating  temperature.  The
     power switch should  be turned on;  however,  it  is not necessary  to  turn
     the pumps  on.  Also,  the unit  should be run  for at  least 15 minutes with
     filters  in  the  sampling position and the front access  door closed.
                                      32

-------
7.    After the instrument is warmed up, the filter seal  system must be checked
     for leaks.   Insert a pair of filters (not the first filter set) and close
     the filter seal  mechanism using the manual  switch.   Turn on the pumps and
     plug the flow through the virtual  impactor by either inserting the rubber
     stopper or by inserting  the calibration adaptor described  in  Appendix I
     and closing  the adaptor  valve.   Flow  through  both the  fine  and coarse
     flowmeters should  drop to  zero.   If any  flow is  indicated,  the filter
     seal mechanism  should  be  checked.   The limit switches on the filter seal
     mechanism  and  the  seal  tube  height  can  be adjusted  following  the
     instructions provided  in Section  6.2.3 and  6.2.4  of  the manufacturer's
     manual (2).

     Another possible  source  of  flow leaks  is  in  the  blocks that support the
     impactor  assembly  (see   Figure  12).  In  later  production  models  these
     supports were machined down 0.03 in. to provide a tighter seal around the
     filters.   Operators of  early models of the  instrument  should consider
     retrofitting  their  samplers   with   the  modified  support  blocks  (after
     contacting the factory) if a flow leak  is  indicated.

     If  the seal mechanism is  adjusted  properly and  there are no  loose or
     damaged   fittings   or  tubing   but   leaks   are   still   present,   the
     manufacturer's  service   department  should  be  contacted  for  further
     troubleshooting.   If  no  leaks   are indicated  but the  flow  cannot be
     adjusted to obtain 15  liters per minute  through the fine flowmeter, there
     may  be a  problem  with the  flow  controller  or the pump,  and again the
     manufacturer should  be contacted for further  instructions.

8.   Once   the  system  has  been  checked for  flow  leaks  and  adjusted,  if
     necessary,  calibration  of  the  fine and  coarse   flow  system  should be
     accomplished  using  one  of   the  methods  described   below.   The   flow
     measuring   devices  selected  for   the  calibration   must  be  suitably
     calibrated themselves  (e.g.,  standard NBS bubble meter for  low flow, and
     lab-type   calibrated  spirometer  for  high  flow   measurement  equipment)
     before being used  to calibrate the  dichotomous  sampler.
                                     33

-------
     NOTE:    Since  flow  measurements  are  highly  sensitive  to  temperature
     conditions,  equipment used  in  calibrations  and the sampler  should  be  at
     comparable   temperatures    during   calibration.     Allow   equipment   to
     equilibrate   in  the  sampler operating  environment for  about  two  hours
     prior to proceeding with the calibrations.

INSTRUMENT CALIBRATIONS

     Instructions are  provided below  for two methods of fine  and coarse flow
calibration  for  the   dichotomous   sampler:    The  first   uses  ball-in-tube
rotameters, and  the  second  uses  a Singer Model  DTM-115-3 dry test meter.  For
the rotameter  calibration system,  flow calibrations are to be made at a point
above the  filters (referred to  as  "at  inlet")  and a point  below the filters
("at base").   This  will  assist  in detection  of leaks  in  the  filter sealing
mechanism.   It j_s important  to note that leaks HI the filter seal area cannot
be  detected by_  the sampler's built-in flowmeters.  For  the  dry  test  meter
method,  flow calibration is  made only at the inlet because  it is not practical
to perform flow checks both  above and below the filters. If the flow check and
inlet flow  calibration are  done  according to the  instructions  for the Singer
meter, then adequate leak detection will have been accomplished.

     Descriptions,  specifications,  and  supply  sources  for  equipment used in
these procedures are presented in Appendix I.

CALIBRATIONS USING ROTAMETERS

     It  is recommended  that  Types 603  and  604  rotameters as  described in
Appendix  I be used for  calibrations.  As with  all  rotameters,  they must be
corrected  to  temperature and pressure  conditions at  the sampling site.   This
is  most  important when the  rotameters  are  being used outside  in cold weather
and/or  high   altitudes.   For   the  603  and  604  rotameters,  pressure  and
temperature  correction  tables  are  provided  in  Appendix   II;  if  other flow
meters are  used,  see that manufacturer's recommended procedures.
                                      34

-------
     The operator is cautioned to review Item Number 6 under "Instructions for
Sampler Checkout"  to ensure  the instrument is adequately warmed  up prior to
calibration.  Also,  the instrument  must be checked  for leaks  in accordance
with Item 7 of the checkout instructions before proceeding.

Caution:  To avoid possible damage to the pumps do not operate the sampler for
more than 20 minutes with the valves closed.

Fine Flow Calibration (at Base)

1.   Attach the  604  rotameter to the polyethylene  vacuum  tube at the base of
     the fine flow  tube on the side of the filter change mechanism closest to
     the front  access  door,  as  shown  in  Figure 14.  This  is  accomplished by
     loosening the  Swagelock  fitting at the 90° elbow, removing the tube from
     the elbow,  and attaching  it to the  3/8" adaptor  fitting (described in
     Appendix  I).    The  adaptor  is  then  inserted  into  the  heavy  walled V
     vacuum tubing,  and the tubing is attached to the 604 rotameter.

2.   Once  the  tubing  has  been  attached,  the  rotameter   leveled,   and  the
     operator  is in a  position to view  easily both  the  604  and  fine flow
     rotameters, the sampler pumps should be turned on.

3.   Adjust the  fine flow control valve (see Figure 20) to obtain air flow of
     15 jfcpm on  the  604 rotameter usiing the  calibration  curve provided with
     the 604  meter.   Allow the system to run at this flow rate  for 5 minutes.

4.   Record  the  observed  flow  readings  of  both  the  604  and  fine  flow
     rotameters, and of the vacuum gauge.

5.   Using  the  fine flow  control  valve,  adjust  the  flow  through  the  604
     rotameter  to obtain  readings  for  two flow rates  above  15  £pm  and two
     below.  Record  indicated flow readings for both the 604 and the fine flow
     rotameters  for  each point.   Also record the vacuum gauge reading at each
     point.   Using  the 604 rotameter  calibration  curve,  determine  true air
                                     35

-------
                                                                 604 ROTAMETER
                                                              1/4" VACUUM TUSE
                                                             FINE FLOW ROTAMETER
       FILTER OVERLOAD
       ADJUSTMENT SCREW
Figure 14.  Typical Installation for Fine Flow Calibration at Base
            Using 604 Rotameter
                                   36

-------
     flow for each  point,  and plot the indicated versus actual flows.  Figure
     17(a) shows a sample form that will simplify data recording and plotting.
     Separate  forms  are   necessary   for  the  fine   and   the   coarse  flow
     calibrations.

6.    Draw the calibration  curve  and determine the proper  fine flow rotameter
     set point to obtain an actual flow rate of 15 £pm.

7.    Return the  3/8"  polyethylene tube to the 90° elbow Swagelock fitting and
     tighten securely.

Fine Flow Calibration (at Inlet)

1.    Attach  the  604  rotameter  to  the  valve  of  the   inlet  adaptor  tube
     (described  in  Appendix  I)  with the heavy wall vacuum tubing, making sure
     that the valve is  fully open (counter-clockwise), as shown in Figure 15.

2.    Insert and seal a clean set of filter (not filter set #1) into the filter
     mechanism.

3.    Turn the  coarse flow  control  valve (Figure 21)  off  (clockwise) so that
     only fine flow is being drawn through the inlet.

4.    Adjust the  fine flow  rotameter  to  read  the 15  £pm  set  point which was
     determined  at  the  base  calibraiton.  If the 604  rotameter  does  not give
     the  same  reading   as  it  did  during  the  base  calibration,   the  seal
     mechanism is suspect  of a leak and may need adjustment [see Sec. 6.2.3-4
     of Beckman Manual (2)].

5.    If  the  same reading  is  found, proceed to adjust  the  fine  flow  valve to
     obtain two  points  above  15  £pm  and two points  below as  before.  Record
     these  data  including   the   vacuum  gauge  reading  on  the  fine  flow
     calibration form in the "At Inlet" section, Figure 17(a).
                                     37

-------
                                                ADAPTOR FLOW
                                                CONTROL VALVE
                                                   CALIBRATION ADAPTOR
                                                        ROTAMETER
Figure
       15.
                                     Fine
                                          Flow Calibration at
            Adaptor.
                              38

-------
                                                                   603 ROTAMETER
                                                                  TYGON VACUUM TUBE
                                                              COARSE FLOW ROTAMETER
Figure 16.  Typical Installation for Coarse Flow Calibration at Base
            Using 603 Rotameter
                                      39

-------
6.    Again plot  the points which  should be  similar to the  "At  Base"  curve.
     From both curves determine the proper set point.

Coarse Flow Calibration (at Base)

1.    Attach the  603  rotameter  to the "Tygon  type"  vacuum  tube at the base of
     the  coarse  flow tube, the  side of the  filter  change mechanism farthest
     from  the access door  as  shown  in  Figure 16.   This  is  accomplished by
     pulling the Tygon tubing off the nipple at the 90° elbow and attaching it
     to  the  adaptor nipple (See Appendix  I  for  adaptor  description).   This
     adaptor  is  then attached  to the  heavy  wall  vacuum  tubing and  the 603
     rotameter.  Be sure that the Tygon tubing is  not pinched or bent sharply.

2.    Once  the tubing has been  attached  and the  operator  is  in a comfortable
     position  to read both the  603 rotameter and  the  coarse flow rotameter,
     the  sampler's pumps should be turned on.

3.    Adjust  the  coarse  flow valve to  obtain  a reading of 1.7  2pm  on the 603
     rotameter.  Allow the pump to run for 5 minutes.

4.    Adjust  the  valve  to obtain readings for two points above 1.7 £pm and two
     points  below.  Record this  data on  the  coarse  flow  dichotomous sampler
     calibration form Figure 17(b) in the "At Base"  section.

5.    Plot the curve and determine the proper set point to obtain a flow rate
     of  1.7 £pm.

6.    Return  the "Tygon  type"  tubing to  the  90° elbow  and  check to see  that
     both ends of the tube are firmly attached.
                                      40

-------
FINE FLOW DICHOTOHOUS SAMPLER CALIBRATION
                                      Site:
                                                           Date:
DICHOT. CALIBRATOR
TYPE
Ser. No. Ser. No. '
1 	 — At
Last Calib.... Last Calib. Bas
Current Temp, Temp.
Current Press. Atm. Press.
By
Low flow cut out switch set Ij^i
at 1pm 	
Remarks:
_i I
i ~ ~ i i :: 1 1 ::...:: - - T
r. z. - - ...-4
t ~ : . . _ . 4
i _i




,
is • • • : ".'.".'.'.'.'. 	 !; r r "

L tl L [

I ft
u L 1. L
r [Li
L! t i
r r 1
17 L L L

p .£" Jl ' L 1 L I
"• r" "L t- t t t
rH L II I
1 J I
1 1 J 1
I I
5 1 t
5 	 . _ - ~ ± - -i _j -"i ~| - •• r
CM 16 	 ~*~ "h
j
fc __._._-_ .JJ Ii- - ... - . ..]....

< i : i : : i : 1 1 1 1 z 1 1 " 1 1 1 : i : : i j; " * : :
± : 	 : : :: :. ._...._ T

o T 	 " 	 T~
4
t* -m F jT

T
I
i : : : 1 1 • : 	
	 : i i . . ; t i ~ : "
T
: in 	 j — • • r 	
	 - 	 • • • j 	
;;:••••• 	 3 	 •
u :: ; ;. 1 1 :i i : i ii 1 1 ii i Ii 	



~





13 	 " " 	




j j .. i . . 	 	 i
j d . J 	 .1

j Ij ~l
j "1 ^
J -J L
12 (35) (40)
12 13 14 15
Dichot. 604 Rotameter Actual VAC.
Rotameter Indicated lorn ins.
nd lea ted
E



= t

. . - -i- . . 4- •'•
-»

. . J | " "










- h h F

t L L t L—
u. L EZ t [Z
I " r ' L L.
1 1 1 [
1 [- -t I

r Li

1 i
I j~*
i i j~
I r"
T r
t { - ' * ' t " ? - 	 ~








T~
^T~












i




"^






I
i L



t "J "T -p - • -J- - - —
f«5) (50)
16 17 18
                               Observed Flow  Meter Reading
            Figure 17a.   Suggested  Fine Flow Calibration Form
                                     41

-------
COARSE FLOW DICHOTOMOUS SAMPLER CALIBRATION    Site:
                                                           Date:
DICHOT. CALIBRATOR
TYPE

Ser. No, »er. no.
Last. Calib. Last Calib. las
Curr. Temp. Tenp.
Current Press Attn. Press. i

Bv At
Inl
Remarks :
• : : i.. : 	 }..
: _ i 	 4.
:: i. : ------- 4-






it
. : : ± . .
. . : ± : ....
. : : : ±
T





	
§i
. ± 	
1
t
. .1. ..... .
. 4- -• 4-1
, _ .1. . j .1
_L _,_ j~ 1 _ _ _
in -- L. . , a- T -
^ 	 ~ " " il " IL |" " "

< " - r • • P P " t p i" "-" .

"""i " " b b ' " " p - • - -
o "t t t 	 " h
* : - : : ± 	 J ± _ t

1.7 • : : ' : 	

j . . . ......... —
] 	
- - - -_\ 	 	
	 .----.. 	
- • - - i-t-i- 	 - - -


IS . - ...--..-_...-.-..

"
1 t" 1 I
31 . J~ J-
	 	 : . . ± : ± 1
	 i . ± i

1 r
I -L
i ^ _j_
i-4 (14) (15) (16) (17)
1.4 1.5 1-6^ ,1-7
Dichot. 603 Rotameter Actual
Rotameter Indicated 1pm
Indicated
e

et

. 	 . . .t 	 - i -
L
T " " T ~ ~
. ....:.. 	 ± : : : ± : ~




r
	 L . - ... ± . . L .
. . .. ± 	 	 	 	 ± . . r _
. ... .L . . 	 ± . J- -
...L 	 ±.±__
. . . L . 	 -L . 1 - p

• _1_ f" |
I L L
_r~ p
4 ' £ n
i




I * ____, 	
....:., -±-
til - - -- .- 	
1 -1 1.-.: L:..:L, •>: ; L ::::
i 	 	 r- - p ----
• i j i
[ L. J
[ t-
_i r n
L E- J
I t_ n
1 \- -J
i I L _ T
J J i
	 : - tl - - - 	
F :::::::::
I \-
j! iill |! 111
T :: I : i \"~
-- 	 - "










U . - L ± . ..L 	
; . 	 t ± . _ _t 	
j-'LJ' JL JlJI IE-_
(18) (19) (20)
1.8 1.9 2.0
                              Observed Flow Moter Reading
         Figure 17b.    Suggested Coarse  Flow Calibration Form
                                        42

-------
Coarse Flow Calibration (Inlet)

This  procedure  is  similar to  the fine  flow rotameter  (inlet)  calibration.

1.   Attach the 603  rotameter to the valve of the inlet adaptor tube with the
     heavy  wall   vacuum tubing,  making  sure  that  the valve  is  fully  open
     (counter-clockwise).

2.   Insert and  seal a clean  set  of  filters (not filter #1)  into the filter
     mechanism.

3.   Turn  the  fine  flow  control valve  (Figure  21)  fully  off (clockwise) so
     that only coarse flow is being drawn through the inlet.

4.   Adjust the coarse  flow control valve to read the  1.7 £pm set  point which
     was  determined  at  the  base calibration.  If the  603  rotameter does not
     give  the  same  reading  as  it  did during the base calibration, the seal
     mechanism is suspect of a  leak and may  need  adjustment. (See Sec. 6.2.3-4
      in Beckman's Manual).

5.    If the  same  reading is  found, proceed  to  adjust  the valve  to obtain two
     points above 1.7 £pm and  two  points below as before.   Record these data
     on  the fine flow  calibration form in the "At  Inlet"  section [Figure
 6.    Again  plot the  points  which  should  be similar  to the "At Base"  curve.
      From both  curves  determine  the proper  set  point.

 Total  Coarse/Fine  Flow Check  Using  Only  the 604 Rotameter  at the  Sampler Inlet

      After  the complete  calibrations for  both  the  fine flow and coarse  flow
 have  been performed,  a total  flow check  can be  performed as  follows:
                                      43

-------
1.   Attach  the  604 rotameter to the  inlet adaptor and then  to  the inlet of
     the sampler as shown in Figure 15.

2.   With a set of filters (not #1) clamped and sealed, and the pumps running,
     adjust the fine and coarse rotameters to the set points determined by the
     calibration.

3.   Read the 604 rotameter and determine the total flow rate.

4.   Next turn the  coarse  flow valve  (fully clockwise)  off.   Observe the 604
     rotameter reading  as  the  fine flow. The difference between the total and
     fine flow  rotameter reading  should be the coarse  flow.  This  flow check
     should be performed at the time of calibration and weekly, and the values
     recorded.

CALIBRATIONS USING A DRY TEST METER

     By using a Singer Model DTM-115-3 dry test meter connected at the sampler
inlet it  is  very  easy to accurately set the fine and coarse flows.  First, one
closes both  the  fine and coarse  flow  valves to verify  that both valves seat
properly.  Next,  one opens the coarse flow valve and adjusts  it so that the
Singer gauge  and  a  stopwatch indicate an  inlet  flow rate of 1.67 £pm.  Since
the fine flow is still shut, the inlet flow and coarse flow are the same.  Now
adjust the fine  flow to 15 £pm, but do not adjust the coarse flow.   A correct
adjustment of the  fine  flow will  result in a total inlet flow of 16.67 £pm as
indicated by the Singer gauge and stopwatch.  A simple test for leaks can also
be made.   The step-by-step instructions are as follows:

1.   Connect  the  Singer dry  test meter,  as  shown  in  Figure  18.  Securely
     tighten the ferrule to prevent leaks in the calibration system.

     The  following  leak test is only  appropriate  when the aerosol  sampler is
     being  set  up  or  tested.   During  routine  aerosol  sampling,  proceed
     directly to step 3.
                                     44

-------
                    SINGER DRY TEST METER
Figure 18.  Typical  Installation for  Calibrations Using  the  Singer Dry  Test
           Meter
                                     45

-------
2.    a.    While operating  the  sampler  in  the manual  mode,  insert a  pair  of
          filters,  engage the filter seal,  and turn on the pumps.

     b.    Close the  coarse  particle  flow valve  on the  dichotomous  sampler.

     c.    Close the valve at the sampler inlet shown in Figure 18.

     d.    If there are  no  leaks,  then the fine particle flow indicator on the
          dichotomous  sampler   should  drop  to  zero,  and  the  gauge  at  the
          sampler inlet  (see Figure 18) should indicate  a vacuum  of  at least
          23 in. (58 cm) Hg.

     e.    If there appears to be a leak, first verify that the ferrule, valve,
          etc,  on  the  sampler  inlet  (see  Figure 18) are not  leaking.  If the
          leak  is  traced  to the  sampler itself,  then  the  sampler  requires
          repair or adjustment.

3.   Open  the  inlet  valve shown in  Figure  18.   Close both fine  and coarse
     valves  and verify that the  flow is less than 0.1 £pm  by observing that
     the  pointer  on  the  Singer gauge  changes  by  less than 0.1  liter  in 60
     seconds.

     Caution:   To  avoid possible  damage to pumps  do  not operate  the sampler
     for  more  than 20 min. with valve(s) closed.

4.   Open the  coarse  particle flow  valve, and as  an initial approximation, set
     the  coarse flow  to 1.7 £pm  using the  instrument's rotameter as a flow
     indicator.

5.   Measure the time  for the Singer  gauge pointer  to  change  by at least 2
     liters.   For  a coarse particle  flow rate  of 1.67  £pm,  the time should be
     72.0 seconds  for 2 liters.
                                      46

-------
6.    If the measured  time  deviates from 72.0 seconds by more than 1.0 second,
     adjust the  coarse  flow and  measuring  the time  until  a  correct coarse
     particle  flow rate is obtained.

     Verify the final  value with an additional measurement

7.    Now open  the  fine particle flow valve,  and as  an initial approximation,
     set the  flow to  15 £pm using  the instruments fine  rotameter  as a flow
     indicator.

8.    Measure the time for the Singer gauge pointer to change 10 liters.  Since
     a total  flow  of  16.67 £pm should be flowing into the inlet (15 fine plus
     1.67 coarse), the stopwatch time should read 36.0 seconds.

9.    If the measured time deviates from 36.0 seconds by more than 0.4  seconds,
     keep adjusting the  fine flow and measuring the time until a correct total
     flow is obtained.

     Verify the final value with an additional measurement.

FILTER OVERLOAD SENSING  SWITCH SETTING

     The  differential   pressure   switch  which  detects  a   filter   overload
condition is controlled  by using the adjustment screw  shown in  Figure  14. This
screw  has  been  preset  by the  manufacturer  and  should  not  need resetting.
However,  it is recommended that the pressure differential setting be checked
after  completion  of instrument calibration  and prior  to initiating a  sampling
program.  This check  should  also  be  performed as part  of the six-month flow
audit  described in Section 7, Quality Assurance.

Rotameter Method

     To  check the  filter overload cutout  setting,  install  the 604  rotameter
and  flow calibration adaptor.   Turn  off  the coarse flow  completely.    The
                                      47

-------
instrument should be in the automatic mode.   Begin to slowly turn the valve on
the adapter  tube clockwise.   This  will  restrict the flow  to  the dichotomous
sampler  inlet.    Note  that  as  the  flow  is  restricted,  the  sample  will
automatically compensate by  increasing  the  flow to the normal  value.  Closely
observe the 604 rotameter and note that at some point that the flow controller
no longer  can compensate  for the restriction and the total  flow will begin to
decrease.  Closely  watch the  604 rotameter while continuing  to  restrict the
flow with  the valve.   Eventually the cutout switch  will  cause the sampler to
switch to  another  filter.   It is at  this cutoff point that the 604 rotameter
reading  should  be recorded.   Reset  the  sampler in the automatic  mode on the
same filter set.  Repeat the above procedure a total  of three times, recording
the  604  rotameter reading  each time  that the cutoff  switch  switches the
filters.

     The filter overload pressure switch has been adjusted to actuate when the
fine  flow  decreases from 15  £/min  to 14.25 £/min by  filter  loading.   If the
check  indicates  a  significant difference from this setting, the switch should
be adjusted.

     To  do  this,  rotate  center  adjusting  screw  on  switch  (Figure  14)
counter-clockwise  to  end of  travel.   Place instrument  in  automatic  sampling
mode and set fine flow at 15 £/min.    Restrict  inlet flow using the calibration
flow  adaptor  valve shown in  Figure 15 until flow drops to 14.25 £pm,  and then
rotate  the pressure  switch adjusting screw  clockwise  until  the filter pan is
switched.  Back  the screw out h  turn  from this point.  Contact Beckman  if this
method does not work.

Dry Test Meter  Method

1.    Perform  steps  1 and 2 described in the above section, Calibrations  Using
      a Singer Dry Test Meter.
                                      48

-------
2.    Open the inlet flow valve shown in Figure 18.

3.    Start sampling in the automatic mode.

4.    Close the sampler's coarse particle flow valve.

5.    Set  the  flow rate  to 15 £pm  as measured  at the  inlet.   For this a 10
     liters  volume should  pass   through  the  dry  test  meter  in  40 seconds.

     To  determine  how the filter overload has been set previously,  skip steps
6-8, and proceed directly to step 9 below.

6.    Disable the filter overload  switch by_rotating its adjustment  screw fully
     counter clockwise (see Figure 14) to end of  travel.

7.    Partially  close   the  inlet  flow rate restrictor  so that  the flow rate
     decreases  by 7%.   At this  point  the dry  test  meter  should  require 43
     seconds to  pass  10 liters.   Note the  vacuum reading: it  should indicate
     about  5  to  10  in.   (12  to  25 cm) Hg.   This represents that  the maximum
     pressure drop across a filter that will  cause  a  7%  decrease in flow rate.
     If  the vacuum is lower than  the  above  range of  values,  then there is
     probably a  leak  in the flow  system.

8.   Slowly  rotate the  filter  overload  adjustment screw clockwise until  the
     filters are  caused to change.

9.   Verify  the  setting of the filter overload  adjustment screw.    To do this
     first  open  the inlet  flow restrictor valve.   Then  slowly close it so that
     the vacuum rises  by about  1  in.  (2.5  cm)  Hg every 10  seconds.   Verify
     that  the  filters are caused to  change when  the  vacuum  has the same value
     that was set  in  step  7 above.
                                      49

-------
                               SECTION 6

                         OPERATING PROCEDURES

     It is recommended  that  the  operator become familiar with  the  principles
of  operation  (Section  2)  and  the  instructions  below  before  attempting  to
operate the instrument.

     In testing the  sampler  or in learning how to operate it,  it is unwise to
use clean  filters  that  have  been preweighed for  aerosol  sampling.   It is far
better to  use a  pair  of trays  that are  used  only for  practice  or testing.

SAMPLER HOUSING

     The dichotomous  sampler  is  housed in  a  weather  resistant shelter  that
must be firmly in place during sampling.  The front door of the shelter may be
removed by turning the  keylock  clockwise  and  pulling the door  out and down.
This  will   provide  access  to  the  keypad,  printer,  sample  trays and  flow
controls.    Maintenance  and  troubleshooting may  necessitate  removal   of  the
entire shelter.   To do  this,  raise  and remove the inlet  standpipe from the
shelter.    Remove the metal screws around the base of the shelter, and lift the
shelter up until it clears the instrument.

     CAUTION:  Remember that whenever the standpipe has been removed, a No. 1\
or  8  rubber stopper or  other  suitable  device  must be inserted  in  the top of
the impactor to prevent dirt or foreign  objects from entering.
                                      50

-------
FRONT PANEL CONTROLS

     The controls for programming and operating the sampler are located on the
front control panel (Figure 19).   Their functions are described below:

1.    POWER  Switch:   When  button  is  depressed,  power  is   applied  to  the
     instrument; pushbutton is illuminated when power is on.

2.    PAPER ADVANCE  Switch:   Paper in the printer is advanced when this switch
     is depressed.

3.    KEYPAD:  The keypad controls all instrument functions for programming and
     operation.

4.    STATUS  Lights:   8  light   emitting  diodes  (LED's)  which  indicate the
     operational mode of the instrument, and identify malfunctions.

5.    INSTRUCTION  Lights:   8  light emitting  diodes (LED's)  which  help the
     operator set up the instrument.

6.    MANUAL  CONTROL Switches:

     MANUAL  OFF-ON (far  left  switch):   When  ON,  permits  use of 4 remaining
     switches  (below) to test operation manually.

     SHUTTLE MOTOR  (second switch  from  left):   When  depressed,  this switch
     will  drive  the shuttle mechanism motor.

     SEAL  MOTOR (Middle switch):  When depressed,  will  drive the filter  seal
     motor.

     PUMP  MOTOR  (second  switch  from right):  When  on,  will  provide power  to
     both  the  fine  and  coarse flow  pumps.
                                      51

-------
                                STATUS LIGHTS      INSTRUCTION LIGHTS
        PAPER ADVANCE
        SWITCH
POWER SWITCH
                                                                          KEYPAD
              CIRCUIT  BREAKERS
MANUAL CONTROL 'SWITCHES
             Figure 19.  Sampler Front Panel Controls
                                     52

-------
     TRAY MOTOR  (far  right switch):   When  depressed,  will  drive the  tray
     indexing motor.

7.    HEATER,  MAIN  POWER  and  MOTORS  Circuit  Breakers:   Individual  circuit
     breakers are provided  for  the following: main power, heater and electric
     motors.   They   trip  and  remove  AC  power  from  the  equipment when  an
     overload exists.   To reset,  turn  the POWER  switch  off  and  momentarily
     push in  on  the  tripped  circuit breaker switch.   Turn  the  POWER switch
     back on  and make  sure it  is  operating  properly.   If  a circuit breaker
     trips  again,  do not  reset.   This  is an indication  that further repairs
     are needed.

INTERIOR FLOW CONTROLS

     The  dichotomous sampler is  equipped with a  flow meter and  flow control
valve  for each of the  fine and  coarse  flow  circuits, as  shown  in Figure 20.
The  larger  flow  meter and its valve  control  the  fine  flow.  This  flow is
nominally   set  at  15  £pm and  then  carefully  set  with  an accurate  flow
calibrator.    The coarse  flow  (smaller flow  meter) is then set  at  1.7  £pm.

     As  described in Section  5 of this report, a  vacuum pressure  gauge should
be installed to provide additional flow measurements.

OPERATING INSTRUCTIONS

     This section provides detailed, step-by-step  instructions for programming
and  operating the  dichotomous  sampler.   An example of a  typical  sampling
program  is  given along the right-hand edge  of the instructions  to assist in
relating  sampling   requirements  to  instrument  instructions.  A simplified
summary  of  these instructions is given at  the end  of this  section  for use as  a
guide  once  the operator is  familiar with the  instrument.
                                      53

-------
                      COARSE FLOW ROTAMETER
FINE FLOW
CONTROL VALVE
FINE FLOW ROTAMETER
                                                                     CAPS  AT  FORMER
                                                                     LOCATION OF
                                                                     PRESSURE SWITCH
                                                                    COARSE  FLOW
                                                                    CONTROL VALVE
                                           RECOMMENDED INSTALLATION
                                           OF VACUUM GAUGE
     Figure 20.
                  Interior Flow Controls, Showing Recommended Installation
                  of Vacuum Gauge
                                         54

-------
 OPERATION
              INSTRUCTION
EXPLANATION/EXAMPLE
    1.
 Record
 Keeping
 System
Enter appropriate information for logs or
record keeping system to insure that
necessary data for sample collection is
properly supplied and recorded.
   2.
 Turn On
Plug instrument in and depress power
switch.  Immediately, power switch should
light, printer should advance paper
approximately 2", and pump should run
momentarily.  Instruction LED's (light
emitting diodes), CALENDAR TIME and DAY
should light.
   3.
Set
Calendar
Date
Key in on the keyboard a 3-digit number —
001 through 365 -- corresponding to the
present date.  Refer to Figure 21 for
date and corresponding Julian Day to be
used.  DAY will turn off, and HOUR will
light.
           CAUTION:   Operator should note whether
           current year is a leap year, i.e., 1980,
           1984, etc., and use the appropriate Julian
           calendar in Figure 21.  The sampler is
           designed to print Day #1 after Day #365.
           The operator must adjust the dates on the
           printed record when sampling at the end of
           a leap year.  See Figure 21, Part B (Leap
           Year Julian Calendar) for more detailed
           instructions.
Assume today's date is
March 10, 1980.  This
is a leap year, so look
up March 10 on leap
year calendar.
Corresponding Julian
Day is 70.  Key in
070.
   4.
Set
Calendar
Time
Key in a 2-digit number — 00 through 23 --
corresponding to the present hour.  HOUR
will turn off, and MINUTE will  light.
Key in a 2-digit number -- 00 through 59 --
corresponding to the present minute.
MINUTE will turn off, and printer will
print out the present calendar  date and
time.
Assume time at  the
moment is  3:23  pm.
Key  in 15  for HOUR  and
23 for minutes.   Print-
out  will be as  follows:

        DAY  HOUR  MINUTE
#1      U70  ~I5~~    23
                                     55

-------
OPERATION
INSTRUCTION
EXPLANATION/EXAMPLE
           Following this printout, the filter shuttle
           will automatically retract, and the filter
           seal mechanism will open.   The tray INDEX
           ERROR may then light.
  5.        The filters are notched in one corner to
Filter     insure correct positioning and sealing.   As
trays      shown in Figure 8, the proper orientation
           of the filters is with the notched corner
           facing the impactor assembly and away from
           the front of the instrument.   Also, note that
           the fine filter tray is closer to the front
           of the sampler and the coarse filter tray
           is to the rear.  Raise the filter tray
           carriage to its uppermost position using
           the carriage release knob, and insert both
           trays.  Tighten the filter carriage
           guide bar across the top of the filter
           trays.  Make sure the trays fit squarely
           in the tray holders and against the small
           metal tabs in the corner of the tray holders
           (See Figure 8).
  6.       Once the trays are loaded, lower the en-
Tray       tire assembly completely.  Depress RESET
Position   for approximately one second.  If trays
           are positioned and seated properly, the
           INDEX ERROR light will go out, and START
           TIME and DAY will light.  Depress TRAY
           ADV and a two digit number (02 to 36)
           of the desired filter for the start of
           sampling. The first pair of filters must
           always be kept unexposed as a blank.
                              Assume that sampling
                              is to begin at tray
                              position 3.  Depress
                              TRAY ADV and key in
                              03.
  7.       Key in a 3-digit number -- 001 through
Sampling   365 -- for the desired starting date
Start      of the sample.  Refer to correct
Date       Julian calendar (Figure 21).  DAY will
           turn off, and HOUR will light.
                              Assume the desired
                              date for sample
                              collection to start
                              is March 15, 1980.
                              Julian day is 075.
                              Key is 075.
                                  56

-------
    .IUI.1AN CALENDAR SHOWING DATE AND CORKICSKINDINO JULIAN DAY FOR NON-I.FAP YKAR




                      NOTE: CALENDAR ONLY SHOWS KVliRY OTHER DAY
Jan. 1 1
Jan. 3 3
Jan. 5 5
Jan. 7 7
Jan. 9 9
Jan. 11 11
Jan. 13 13
Jan. 15 15
Jan. 17 17
Jan. 19 19
Jan. 21 21
Jan. 23 23
Jan. 25 25
Jan. 27 27
Jan. 29 29
Jan. 31 31
Feb. 2 33
Feb. 4 35
Feb. 6 37
Feb. 8 39
Feb. 10 41
Feb. 12 43
Feb. 14 45
Feb. 16 47
Feb. 18 49
Feb. 20 51
Feb. 22 53
Feb. 24 55
Feb. 26 57
Feb. 28 59
Mar. 2 61.
Mar. 4 63
Mar. 6 65
Mar. 8 67
Mar. 10 69
Mar. 12 71
Mar. 14 73
Mar. 16 75
Mar. 18 77
Mar. 20 79
Mar. 22 81
Mar. 24 83
Mar. 26 85
Mar. 28 87
Mar. 30 89
Apr. 1 91
Apr. 3 93
Apr. 5 95
Apr. 7 97
Apr. 9 99
Apr. 11 101
Apr. 13 103
Apr. 15 105
Apr. 17 107
Apr. 19 109
Apr. 21 111
Apr. 23 113
Apr. 25 115
Apr. 27 117
Apr. 29 119
May 1 121
May 3 123
May 5 125
May 7 127
May 9 129
May 11 131
May 13 133
May 15 135
May 17 137
May 19 139
May 21 141
May 23 143
May 25 145
May 27 147
May 29 149
May 31 151
June 2 153
June 4 155
June 6 157
June 8 159
June 10 161
June 12 163
June 14 165
June 16 167
June 18 169
June 20 171
June 22 173
June 24 175
June 26 177
June 28 179
June 30 181
July 2 183
July 4 185
July 6 187
July 8 189
July 10 191
July 12 193
July 14 195
July 16 197
July 18 199
July 20 201
July 22 203
July 24 205
July 26 207
July 28 209
July 30 211
Aug. 1 213
Aug. 3 215
Aug. 5 217
Aug. 7 219
Aug. 9 221
Aug. 11 223
Aug. 13 225
Aug. 15 227
Aug. 17 229
Aug. 19 231
Aug. 21 233
Aug. 23 235
Aug. 25 .237
Aug. 27 239
Aug. 29 241
Aug. 31 243
Sept. 2 245
Sept. 4 247
Sept. 6 249
Sept. 8 251
Sept. 10 253
Sept. 12 255
Sept. 14 257
Sept. 16 259
Sept. 18 261
Sept. 20 263
Sept. 22 265
Sept. 24 267
Sept. 26 269
Sept. 28 271
Sept. 30 273
Oct. 2 275
Oct. 4 277
Oct. 6 279
Oct. 8 281
Oct. 10 283
Oct. 12 285
Oct. 14 287
Oct. 16 289
Oct. 18 291
Oct. 20 293
Oct. 22 295
Oct. 24 297
Oct. 26 299
Oct. 28 301
Oct. 30 303
Nov. 1 305
Nov. 3 307
Nov. 5 309
Nov. 7 311
Nov. 9 313
Nov. 11 315
Nov. 13- 317
Nov. 15 319
Nov. 17 321.
Nov. 19 323
Nov. 21 325
Nov. 23 327
Nov. 25 329
Nov. 27 331
Nov. 29 333
Dec. 1 335
Dec. 3 337
Dec. 5 339
Dec. 7 341
Dec. 9 343
Dec. 11 345
Dec. 13 347
Dec. 15 349
Dec. 17 351
Dec. 19 353
Dec. 21 355
Dec. 23 357
Dec. 25 359
Dec. 27 361
Dec. 29 363
Dec. 31 365

Figure  2la     Julian  Calendar  for  Non-Leap Year

-------
Ln
oo
                  JULIAN CAJ.KNDAK SHOW INC DAIT. AND JULIAN DAY FOR UiAPjnJAR_ (I. c-.^ 1980, J984_, etc._)

                                  NOTK:  CALENDAR SHOWS ONLY EVF.RY OTHER DAY
.Ian. 1 1
Jan. 3 3
.Inn. 5 5
Jan. 7 7
.Inn. 9 9
.Ian. 11 11
.Ian. 1 ) 1J
.Jan. 15 15
Jan. I/ 17
Jan. 19 19
.Inn. 21 21
Jan. 23 23
Jan. 25 25
Jan. 27 27
Jan. 29 29
Jan. 31 31
Feb. 2 33
Feb. 4 35
Feb. 6 37
Feb. 8 39
Feb. 10 41
Feb. 12 43
Feb. 14 45
Feb. 1ft 47
Feb. 18 49
Feb. 20 51
Fob. 22 S3
Feb. 24 55
Feb. 2h 57
Feb. 28 59
Mar. 2 62
Mar. 4 64
Mar. 6 66
Mar. 8 68
Mar. 10 70
Mar. 12 72
Mar. 14 74
Mar. 16 76
Mar. 18 78
Mar. 20 80
Mar. 22 82
Mar. 24 84
Mar. 26 86
Mar. 28 88
Mar. 30 90
Apr. 1 92
Apr. ) 94
Apr. 5 96
Apr. 7 98
Apr. 9 100
Apr. H 102
Apr. 13 104
Apr. 15 106
Apr. 17 108
Apr. 19 110
Apr. 21 112
Apr. 23 114
Apr. 25 116
Apr. 27 118
Apr. 29 120
May 1 122
May 3 124
May 5 126
May 7 128
May 9 1 10
May 11 132
May 13 1 34
May 15 136
May 17 138
May 19 140
May 21 142
May 23 144
May 25 146
May 27 148
May 29 1 50
May 31 152
June 2 154
June 4 156
June 6 158
June 8 160
June 10 162
June 12 164
June 14 166
June 16 168
June 18 170
June 20 172
June 22 174
June 24 176
June 26 178
June 28 180
June 30 182
July 2 184
July 4 186
July 6 188
July 8 190
July 10 192
July 12 J94
July 14 196
July 16 198
July 18 :uO
July 20 202
July 22 204
July 24 206
July 26 208
July 28 210
July 30 212
Aug. 1 214
Aug. 3 216
Aug. 5 218
Aug. 7 220
Aug. 9 222
Aug. 11 224
Aug. 13 226
Aug. 15 228
Aug. 17 230
Aug. 19 232
Aug. 21 234
Aug. 23 236
Aug. 25 238
Aug. 27 240
Aug. 29 242
Aug. 31 244
Sept. 2 246
Sept, 4 248
Sept. 6 250
Sept. 8 252
Sept. 10 254
Sept. 12 256
Sept. 14 258
Sept. 16 260
Sept. 18 262
Sept. 20 264
Sept. 22 266
Sept. 24 268
Sept. 26 270
Sept. 28 272
Sept. 30 274
Oct. 2 276
Oct. 4 278
Oct. 6 280
Oct. 8 282
Oct. 10 284
Oct. 12 286
Oct. 14 288
Oct. 16 290
Oct. 18 292
Oct. 20 294
Oct. 22 296
Oct. 24 298
Oct. 26 300
Oct. 28 302
Oct. 30 304
Nov. 1 306
Nov. 3 308
Nov. 5 310
Nov. 7 312
Nov. 9 314
Nov. 11 316
Nov. 13 318
Nov. 15 320
Nov. 17 322
Nov. 19 324
Nov. 21 326
Nov. 2 1 328
Nov. 25 BO
Nov. 27 332
Nov. 29 3)4
Dec. 1 336
Dec. 3 338
Dec. 5 340
Dec. 7 342
Dec. 9 344
Dec. 1) 346
Dec. J3 348
Dec. 15 350
Dec. 17 352
Dec. 19 354
Dec. 21 356
Dec. 23 358
Dec. 25 360
Dec. 27 362
Dec. 29 364
Dec. 31 366

    Figure 21b.   Julian Calendar  for Leap Years

    Note:  The control system for  the Beckman Sampler is  designed to accept only  Julian days up  to  365.  If
          sampling is to occur  on  Dec.  31 of a Leap Year (i.e.,day nr. 366 in  1980,  1984, etc.),the  sampler
          will record this  day  as  001 of the following  year.  The operator must be aware of  this and  manually
          correct the printer output for day 366 and 001 and then at a convenient point after January 1, the
          printer can be set one  day back to coincide with the normal non-leap year Julian  Calendar.

-------
OPERATION
                  INSTRUCTION
 EXPLANATION/EXAMPLE
      8.
    Set
    Sampling
    Start
    Time
    Key  in  a  2-digit  number  --  00 through
    23 -- for the  desired  starting  hour  of
    of the  sample.  HOUR will  light.   Key
    in a 2-digit number — 00  through  59 --
    for  the desired starting minute of the
    sample.   START TIME and  MINUTE  will  turn
    off, and  SAMPLE DURATION and DAY will
    light.
 Assume desired time
 for sample collection
 to start is 2:30 p.m.
 Key in 14 for HOUR
 aand 30 for MINUTE.
      9.
    Set
    Sampling
    Duration
    Time
             A.
    The  following three  operations  will  set
    the  total  time in days,  hours and  minutes
    that the pump will  run for each sample
    collected.
    Key in a 3-digit number —  000 through
    365 -- for the number of days  (24-hour
    increments) over which each sample is
    to be collected.   DAY will  turn off and
    HOUR will  light.
Assume desired dura-
tion of each collection
is 30 hours. This is
1 day plus 6 addi-
itional hours. Key in
001 for DAY.
                 Key in a 2-digit number -- 00 through
                 23 -- for number of additional hours
                 over which sample is to be collected.
                 HOUR will turn off, and MINUTE will
                 light.
                                              Key in 06 for 6
                                              additional  hours over
                                              first day.
                 Key in a 2-digit number -- 00 through
                 59 -- for the number of minutes over
                 which the sample is to be collected.
                 MINUTE will turn off and DELAY TIME
                 and DAY will light.
                                              Key in 00 for minutes,
                                              as no additional  minutes
                                              are desired for sample
                                              collection.
     10.
   Set
   Filter
   Delay
   Time
The following three operations will set
the total time in days, hours and minutes
between each sample collection.   This
time ^s always measured from the beginning,
or start time, of one group of filters to
tfie beginning, or start time, of the nexl
group, and is called the DelayHMme.
                                         59

-------
OPERATION
              INSTRUCTION
EXPLANATION/EXAMPLE
10. Cont.
Set
Filter
Delay
Time
A.  Key in a 3-digit number -- 000 through
    364 — for the desired number of days
    of delay between the sampling start
    of each group of filters.  DAY will
    turn off, and HOUR will light.
    Still using our example,
    Start Time was set for
    2:30 p.m. on March 15.
    Six days from March 15
    means automatic sampling
    will begin again on
    March 21 at 2:30 p.m. Key
    in 006 for filter group
    delay time of six days
    from beginning of sample
    to beginning of next
    sample.
   Assume sample
   collection is
   desired every six
   days.  This means
   six days from the
   Tast group of filters
   to the start time of
   the next set of
   TTTters.
                  Key  in a 2-digit number -- 00 through
                  23 -- for the desired  number of  hours
                  of additional time delay between the
                  sampling start of each group of
                  filters.  HOUR will  turn off, and
                  MINUTE will  light.
                                              Since no additional
                                              hours are desired
                                              over and above the
                                              six day delay time,
                                              key in 00.
                  Key  in  a  2-digit  number  --  00  through
                  59 -- for the  desired  number of  minutes
                  of additional  time  delay between the
                  sampling  start of each group of
                  filters.   MINUTE  will  turn  off,  and
                  GROUP FILTER COUNT  will  light.

                  NOTE:   If a delay time of 000  days,
                  00  hours  and 00 minutes  is  keyed in,
                  the  instrument will automatically
                  switch  into a  continuous mode,
                  sampling  consecutively through all
                  35  filter pairs.
                                               Since  no  additional
                                               minutes are  desired
                                               over and  above  the
                                               six day delay time,
                                               key in 00.
                                     60

-------
OPERATION
                  INSTRUCTION
EXPLANATION/EXAMPLE
      11.
    Set
    Filters
    per Group
    Number
    Key in a 2-digit number -- 00 through 36 -•
    for the desired number of filter pairs to
    be sampled at each sampling time.   Note
    that each sample actually involves
    collection for a filter pair, one coarse
    and one fine.
   Assume 2 filter
   pairs are desired
   for each collection,
   2 fine and 2 coarse.
   Since our sampling
   duration has been
   set at 30 hours,
   sampling time for
   for each filter
   pair will be 30
   hours.  Key in 02
   for 2 filter
   pairs per group.
                                                   DAY HOUR MIN  DATE
                                                   #2  075  14   30  START TIME
                                                                 00  SAMPLING TIME
                                                                 00  DELAY TIME
                                                                 02  FILTER PAIR
                                                                   per GROUP
                                      #3  001  06
                                      #4  006  00
                                      #5
  12.
Veri fy
Calendar
Time vs.
Start
Time
Push CLOCK PRINT key to verify that the
selected Start Date/Time is later than the
Calendar Time.  If the Start Time is  earlier
than the Calendar Time, go back to operation
#6, and proceed again.

Leave all INSTRUCTION LED's off and
STAND BY on.   The instrument will
automatically wait for the Calendar
Time to coincide wth the Start Time,
at which time sampling will begin
automatically according to the programmed
Still using our example
of today's date, being
March 10, 1980, or 070
on the leap year
Julian Calendar.
Assume Start Date/Time
had mistakenly been
entered as March 5,
1980, or 065 on the
leap year Julian
calendar.  Lower tray,
depress RESET.  Depress
TRAY ADV, key in
starting position
and re-enter correct
Start Date, which is
075, and remaining
data in Steps #2
through #5.
                                        61

-------
 SIMPLIFIED OPERATING INSTRUCTIONS
1.    RECORD KEEPING -
Enter  appropriate  data  on  all  required  record
keeping forms.
2.    TURN ON -
Plug  in  instrument  and  turn  on  power  switch.
3.   SET CALENDAR DATE
     AND TIME -
Key in a three-digit DATE number - 000
through 365,  a  two-digit HOUR number - 00 through
23,  and a  two-digit  MINUTE  number -  00 through
59  -  corresponding  to  present  date  and  time.
Refer  to  appropriate  Julian  Calendar  for  DATE
entry.
4.   LOAD FILTER TRAYS -
Properly   insert   filter   trays.    Lower  trays
completely.  Press  RESET for one  second.
     TRAY ADV -
Press  TRAY  ADV to  advance  tray  to  filter pair
desired (not first pair).
     SET SAMPLING START
     DATE AND TIME
Key  in  a  three-digit  DATE number  -  000 through
365,  a two-digit HOUR number - 00 through 23, and
a  two-digit  MINUTE  number  -  00  through 59, for
desired  sampling  start  date and  time.   Refer  to
appropriate  Julian Calendar for DATE entry.
      SET  SAMPLING  DURATION   Key  in  a  three-digit  number  - 000 through
      TIME                   365,  a  two-digit number  -  00  through  23,  and  a
                             two-digit number -  00 through 59,  for the  total
                             number  of  days,  hours  and minutes  of  sampling
                             duration  for  each sample  collected.
                                       62

-------
8.    SET FILTER DELAY       Key in a three-digit number - 000 through
     TIME                   365,  a  two-digit  number  -  00  through 23,  and  a
                            two-digit number  - 00  through  59, for  the total
                            number of days, hours and minutes of desired delay
                            time  between  the  sampling  start  of  each filter
                            group.   Inputs  of  zero days,  hours  and minutes
                            automatically  initiate  the  continuous  sampling
                            mode.

9.    SET FILTERS PER        Key  in  a  two-digit  number  -  00  through  36 for
     GROUP NUMBER -         the  number  of filter pairs  to  be  sampled at each
                            sampling time.   If zero delay time is selected in
                            Step 7, the Filters per Group number is 00.

     Following  Step #9  (or  #8 if  delay time  is 000  00  00),  printer will
     automatically print out data which has been entered as  follows:

                   (DAY)    (HOUR)     (MIN)
          n        XXX       XX        XX        (Start Time/Date)
          #3        XXX       XX        XX        (Sampling  Time)
          #4        XXX       XX        XX        (Delay Time)
          #5                            XX        (Filters per Group)

9.   VERIFY CALENDAR TIME   Depress  PRINT CLOCK key to  see  if Start  Date/Time
     VS. START TIME -       are   later   than   Calendar   Date/Time.    If  not
                            correct, push RESET and repeat  Steps 4-7.

     STAND-BY  will  light,  indicating  sampler is  in  a "wait" mode  and will
     automatically  begin   sampling  when  Calendar  Date/Time  Reaches   Start
     Date/Time.
                                       63

-------
SAMPLER OPERATION

     Once the  operating  instructions  have been entered  into  the  instrument's
microprocessor,  the  sampler's  clock   runs   until   the  preselected  sample
date/time is  reached.   At  that  point the  sampler starts automatic  control.
The  initial  setup  data  are  printed,   followed by  the  shuttle  mechanism
inserting  a  filter  pair,  the seal  mechanisms  closing,  and  the flow  pump
turning  on.  Filter number  and pumping start time are  printed automatically.

Modes of Operation

     The dichotomous sampler can be programmed for one of two operating modes:
Group  Mode  or  Continuous  Mode.   Group  Mode is  selected with  the  sampling
program  calls for  one  or  more samples  to  be  collected  and then  for  some
interval to pass  before  another sample or  group  of  samples  is collected. For
example, this  mode would be used to collect a 24-hour sample every sixth day.
The  instrument would  be  programmed with a FILTER GROUP COUNT of 1, a SAMPLING
TIME of  1  day,  and a GROUP DELAY TIME of 6 days.   Similarly, if you wanted to
collect  a group  of three samples every 30  days,  the FILTER GROUP COUNT would
be 3,  SAMPLING  TIME  of 1 day, and  a  GROUP DELAY TIME of  30  days.  Note that
the  DELAY  TIME is  the time  between  the  start of each  filter  group,  not the
time between the completion of one group and the start of the next.  Following
completion  of  the last sample of each group, the filter pair will be returned
to the tray  and  the  instrument will  go to  STANDBY until completion  of the
remainder of the GROUP DELAY TIME.

     Although  the Group  Mode  is  considered  the normal  operating  mode, the
instrument  can run in the Continuous Mode as well.  This mode is automatically
selected if the GROUP DELAY TIME is set at zero during initial setup.  In this
mode the sampler  will sequence through all 35 available filter pairs, pumping
for  the  preselected  SAMPLE TIME for  each  pair,  and  automatically changing to
the  next unexposed filter pair as soon as the SAMPLE TIME  is completed.
                                     64

-------
     There is a  disadvantage  in using the Continuous  Mode.   In this mode the
25 seconds needed  to  change filters are not included in any of the selectable
time intervals,  which  will  cause the start times for a sequence of filters to
lag increasingly.  If  one  wants to sample continuously,  on  a fixed schedule,
then use  the  Group Mode,  and set  the  sampling  time to be 1  minute less than
the desired  time between  filter changes.   The  following are  some examples:

Example 1:  Continuous sampling with filter changes every midnight starting on
day 160.

       Day       Hour      Minute
       160       00        00             Start Time
       000       23        59             Sampling Time
       001       00        00             Delay Time
                           01             Filters per group
Example 2:  Continuous sampling with filter changes at 7:00 a.m. and 7:00 p.m.
starting on day 80:

       Day       Hour      Minute
       080       07        00             Start Time
       000       11        59             Sampling Time
       000       12        00             Delay Time
                           01             Filters per group

Filter Overload

     The pump will continue to run for the programmed sampling duration unless
a  filter  overload occurs.   This is  determined  by the  differential  pressure
switch.  If the fine particulate flow rate drops by more than .75 £/min (5% of
the  normal  flow  rate  of 15 £pm),  a signal  from  the  switch  terminates the
pumping  cycle.   The  filter  pair is  returned to the  tray, and  the  calendar
date/time  is  printed on  the  printing tape.   A symbol "v"  is  also printed to
identify the  overload (refer to Figure 22). The next filter pair is inserted,
and pumping resumes.   The pumping time for this second  set of filters will be
                                     65

-------




















CODE DAY HR MIN
»o 1 9 R 1 n n 7 -« .
t 1 t O 1 U U j
»^ nnn n < n n -«--
J UUU UOUU *^~-
»L nni nnnn-«
B^ n 9 «
j V C ^
B1 128 1 Q Q Q -• -.
« 2 128 1003]
» 3 000 0600
» 4 001 0 0 0 0 '
» 5 0 2J

FILTER"
WU. JJAY HR MIN CODE
01 128 1001 S-*-
01 128 1003 T^.
02 128 1003 S
02 128 1003 M^
03 128 1004 S
03 128 1004 *-«-
04 128 1004 S
28 1004 -
128 1005 +-*-
04 128 1006 T
01 128 1008 S
01 128 1008 X-*



PROGRAMMED INPUT
	 	 ..Sampler Start Date/Time
— — — S(imn liner TTTTI^ nr»T Filffti- P-i-i-r
Group Delay Time
- -_ FlIfPT riTnnn Pnnn t~
Calendar Date/Time
Automatic Printout at Sampler Start

OPERATIONAL PRINTOUT



••Pumping Tetmiuate DciLe/Time
	 Fine Filter Already Clogged





Figure 22.  Example of Printer Printout Showing Programmed
            Input Instructions and Operational Outputs
                                    66

-------
the  balance  of  the programmed  sampling  time.   The  sampling time  will  not
include the time  necessary  to switch filters, which is only about 25 seconds.

     If a  second  overload occurs before the end of the sampling duration, the
filters will again  recycle.   Pumping time on the third set of filters will be
the time remaining in the programmed sampling time.

     If  an  overload  occurs  during  Group  Mode  operation,  the  filters  are
recycled, and the follow-on filter pair is counted as one sample of the filter
group count.

Interrupt  Procedure

     The  interrupt  sequence allows the operator to interrupt  a pumping cycle,
change  filter  trays,  and/or  preselect filter,  and  then to continue sampling
with  the existing  instructions  intact.    See  INTPT key operation description
below.

Power  Failure

      In  the event of a  power failure, a  battery will  continue to  supply  power
to the  microprocessor,   and  the  system  wil  continue  in  the  same  mode  and
sequence  as before  loss of  power.   NOTE:  The battery has a capacity of  30
minutes;  if power  is  off for a  longer  period,  the sampler instructions  will
have to  be reset  once  power is restored.

      When all  lights  are  off,   the filter  trays may  be  changed,   since  the
carriage assembly will automatically reset at Filter  #1.   Once the new  tray is
inserted,  the  sampler will continue  with Filter #1 if the CONT  key  is  pushed.
Since Filter #1  must  always be left unexposed,  do not press  CONT.  Instead,
depress  the TRAY ADV  key followed by any two-digit  number 02 through 36 for
the filter number that  sampling is  to continue on.   The  tray will  advance to
the selected filter.   When the tray  stops,  press the CONT key,  and sampling
will continue  in  the original timing sequence.
                                      67

-------
     The pumping on  the  first filter after the interrupt cycle is the balance
of  the  initial  pumping   interval.   The  interrupt  sequence  is noted  by  the
symbol "X" on  the  printer output.  Figure 23 shows a typical interrupt timing
cycle.

Printer Output Format

     The printer tape provides the sampler test history for each sample cycle.
Data  for  each  printout consists of a  code  number or filter number, date/time
data, and sampler function  designator.   Figure  22  shows  a typical printout.

KEYPAD OPERATION

INTPT (Interrupt) Key

      INTPT key interrupts the sampler during automatic cycling and forces the
program  into  a  "wait"  mode.   The  key  is  not  active  during  the automatic
sampler operation.   Depressing the key does  not  affect  initial input  program
data.   Instrument  operation  following actuation of  the  INTPT key depends on
the  operating  mode at  the time the key is depressed.

      Actually,  the  INTPT key during  sampling terminates  the sampling mode,
recycles  the  filters  to  the tray,  and  puts  the  sampler  in  a  "wait" mode.

      Actuating the  INTPT key during  the  filter insertion cycle or during the
seal  closure  cycle  results  in a 5-second pumping  time  followed by the seal
opening  and recylcing  filters to the tray.   The designator "X" and calendar
date/time information  are   printed   at   completion   of   the   pumping   cycle.
Actuating the  INTPT  switch during the  seal  opening  cycle  or during  the  shuttle
return  cycle forces the  instrument  into  the  "wait"  mode  after the  filters are
recycled  to  the  tray.  "X" designator is  not  printed.
                                      68

-------
   « 2   226   1332
   » 3   000   0005
   *4   000   0000
   It 5              36
   0  1
   0  1

   2  0
   20

   2  1
   2  1
  0 1
  0 1

  0 2
  0 2
 226
 226
 1332
 1333
 226
 226
 1337
 1337
  34   226
  34   226

  35   226
  35   226
        1343
        1344
226
226
1348
1353
 S
 X
 226   1335   S
 226   1337   T
 S
 x
        1341   S
        1342   T
          S
          x
226   1344   S
226   1348   T
S
T
                              •Start Date/Time

                              •Sample Time - 5 minutes
                               Zero Delay - Continuous Mode
 Start-Filter  #1
 INTPT
 Tray Advanced to Filter #20
 Start  Pumping #20

 Terminate  Pumping - 5 minutes from
 Start  of Filter  #1

• Start  #21
• INTPT

 Advance Tray to Filter #34
 Start  #34
 Terminate  Pumping - 5 minutes from
 Start  of #21
• Start  #35
 INTPT
 Insert New Filter Tray
 Start  Filter  #1
• Terminate  #1-5 minutes from
 Start  of #35

 Start  #2
 Terminate  #2  after 5 Minutes of
 Pump ing
Figure 23.  Typical Printer Output for Interrupt Cycle
                                     69

-------
PRINT CLOCK Key

     Depressing  the  PRINT  CLOCK key  prints  the  current  calendar  date/time
information.   This key is disabled only during printer operation and is active
at all other times.   A typical clock printout is shown in Figure 23.

RESET Key

     Depressing  the  RESET  key  terminates  the sampling  mode,  recycles  the
filters to the  tray,  and resets all  input  program data.   RESET key is active
at  all  times  after  initially  entering  the  calendar  date/time information.

     Resetting  the  sampler will  cause a TRAY  INDEX  error signal if the RESET
key  is  actuated  while  sampling   filter  pairs  other  than  Pair  Number  1.
Following  filter recycle  the trays  must  be  lowered  to  position 1  and,  the
RESET key must  be depressed before  entering  operating data.

TRAY ADV (Tray  Advance)  Key

     This  key  is used in  conjunction with the  INTPT  key  to  allow selection of
a  specific  filter  pair for  the next sampling position.   NOTE:  The key is
active  only  if the calendar  time is  set, the  trays are in  position 1,  and the
tray  index error light  is  off.

TEST  Key

      The  TEST   Key selects the  instrument  self-test mode.   Depressing  the TEST
 key followed by a numerical key selects  a specific self-test  routine.  This
 key is  active  only  at  power  on,  before   entering  calendar  clock data,  and
 following  the  completion of sample  Pair  Number 36.  The key is inactive during
 sampler operation.
                                      70

-------
CONT (Continue) Key

     This key  restarts  sampler operation under the initial timing parameters.
The  key  is  active  only after  depressing  the INPT   key  and following sample
completion of filter Pair Number 36.
                                     71

-------
                                    SECTION 7

                                QUALITY ASSURANCE

     Throughout this report are  instructions,  recommendations,  and cautionary
notes which  will  greatly  improve  the quality  of data produced,  if  they are
followed  carefully.    In   addition,  the  checks  and  audits  prescribed  below
should be incorporated into the operator's quality control program.

NOTE:  The weekly  flow check  and six-month  flow audit  can be performed using
suitable  calibrated  rotameters,  limited orifice, dry test meters, or similar
flow measuring  devices.   The  instructions below are  for  use  with the 604 and
603  rotamters,  but  similar procedures would be followed in  using the other
instruments.

Weekly Flow Check

     During  periods  of routine operation, the  flow  should be checked and set
as  follows on a weekly basis.  Note:  flow checks should  be made during normal
sampling  operations.

1.   Place the  604 rotameter  and adaptor  assembly on  the  inlet of  the sampler.
     Measure  and  record the  total  flow.  Also  record both the fine and coarse
     rotameter  settings and the vacuum gauge reading.

2.   With the  604  rotameter  in place, turn the coarse flow off.  Record the
     604  rotameter and fine rotameter readings.
                                      72

-------
3.    Turn  the  coarse  flow back  on,  and  adjust  it to  the proper  set point
     determined from the most recent calibration.

4.    Adjust the  fine flow  to  the proper  set point determined  from the most
     recent calibrations.

5.    Again read  the  604 rotameter for total  flow.   If the total flow differs
     by  more  than  0.25 £pm  from 16.7  £pm,  adjust the  fine flow  until  the
     proper  reading  is  obtained on  the 604  rotameter.   Make  sure  that  the
     coarse flow rotameter is still at its proper set point.

6.    Record total  flow  and vacuum gauge readings after final adjustments have
     been  made.   Figure 24  shows a  typical  data log form  for  use  in weekly
     checks.

Weekly Tape Check

     During the  weekly  flow checks described above, the operator should check
the  printer  output  to  see that  the  information  shown matches the programmed
instructions.  The  results  of  this check  should  also be  recorded,  using the
form shown in Figure 24 or other  suitable  log.

Six-month  Flow Audit

     Every  six months  the  coarse, fine,  and total flows  of the dichotomous
sampler  must  be checked  with an  independent calibration  system  using  the
procedure  below. The filter overload cutout  switch  setting should be  checked
at  the same time.    A  604 and 603 rotameter  adaptor tube  assembly other than
the  units  normally used for routine  calibration  purposes,  or some  other type
of  independent  calibration equipment, should  be  used.  All  audit results must
be  recorded  on an  audit log. One copy is  kept on file at the  site,  and one  is
returned  to  the agency which processes  the  audit results.   Figure  25  shows  a
typical  form for recording  this audit information.
                                      73

-------
AUTOMATIC DICHOTOMOUS SAMPLER      AT LAST CALIBRATION:   (AT INLET)

WEEKLY LOG                         CUTOFF SWITCH SET AT (L/M)	
City	      Coarse pts.     Fine pts .      (604) Total pts ,
EPA Site No.	      Roto   Flow     Roto   FLow    Roto    Flow
EPA Serial No.	      	    		
Instrument Mfr.	
Inst. Model No.	      	
Inst. Serial No.                   ^^^^^^^    ^^^^^^^    ^]^^^^
Date of Last                       ^^^^^^_    ^^^^^^^    ^^^^^
   Calibration
Sample Time
   Setting 	      Total/Fine                    Course
Delay  Time 	      Calib  604                    Calib 603
Group Filter Count  	      Ser. No.	           Ser. No.	
Last Cleaning  (Date)               Last Calib.                   Last Calib.	
                                   Atm. Press	           Atm. Press	
                                   By	           By	

Coarse Tray No.	 Fine Tray No.	Tray Type	
Start Date  (Julian Day)
Start Date  (Mo/Da/Yr)_
Start Hour
Fine Flow  (Roto/Flow)
Coarse Flow  (Roto/Flow)
Total Flow (Coarse + FiSeT
Coarse Sample  I.D. No.	
Fine Sample  I.D. No.
 Duration  of Run  (Hrs, Min, Sec)
 Coarse  Sample  I.D. No.	'_
 Fine  Sample I.D. No.
 Duration  of Run  (Hrs, Min,  Sec)
 Sample  Time and  Date Checked
 Against Tape?  (Date + Time)
 Operator's  Initials	Date

 Wkly Flow Check With  604            COMMENTS:
             BFR adj.    AFT  adj.
Tot. flow
Fine flow
Diff.
L/M
L/M
L/M
L/M
L/M
L/M
(coarse)
VAC.
                          Figure 24 Suggested Weekly Log Form

                                      74

-------
    DICHOTOMOUS SAMPLE FLOW RATE AUDIT
    Type  of Sampler 	
                                    Section
                                    Rev.  1 12/79
                                    Page  3 of 3
                                    Form
                                                  Audit Calibrators
    AGENCY

    CITY
                                 Dichot.

                                 SER.  NO.
    SITE  #

    DATE
    AUDITOR
LAST CALIB._

CURR. TEMP._

CURR. PRESS._

BY
Total/Fine
Calib. 604
SER. NO.
LAST CALIF, .
TEMP.
MM. PRESS.
3Y
coarse
Calib. 603
SER. NO.
LAST CALIB.
TEMP .
ATM. PRESS.
BY

    ^OTAMETERS USED AT LAST CALIB.:   SERIAL ft  604
                                                                 X 603
    All tests  to be performed through filters  at  inlet on a normal sampling day.
fVudit Calib. Readi
Total Flow:
504 Roto =
:oarse Off:
504 Roto «
Difference:
Pine Flow :
504 Roto Settings




Coarse Flow:
503 Roto Settings




504
roto: 	 , 	 , 	
ngs



Flow L/Min.










Dichot. Readings
Fine Roto =
Coarse Roto =
Total Flow
Coarse Off:
Fine Roto =
:xi
Fine Roto




Coarse Roto




cut off switch
set at U/M)





Flow L/Min.










Percent Difference



Difference










1.
2.   f:
3.
4.
    The following information pertains' to the sample which was present on the
    monitor during this particular QC inspection:
                               STATION OPERATOR
                                                       OC INSPECTOR
DATE OF SAMPLE
STARTING TIME (MIN. )
ENDING TIME (MIN.)






    COMMENTS:
         Figure  25    Suggested Flow  Rate Audit Form
                                              75

-------
1.    Total flow:   Place the  audit  inlet adapter,  with the  valve  fully open
     (counterclockwise) and the  604  Rotameter attached onto  the  inlet  of the
     dichotomous  sampler  impactor.    Tighten the  hose  clamp and  proceed  as
     follows:

     a.    First, measure the  total  flow as indicated by the 604 rotameter and
          log  it  on the form.   Log  the  dichotomous fine  flow rotameter and
          coarse  flow  rotameter readings  and their  corresponding  flows that
          the  sampler  is  currently  operating at  as  determined by  your last
          calibration.

     b.    Next,  turn the  coarse flow  fully  off  and  log the  604 rotameter
          reading and the fine flow rotameter reading.

2.    Fine flow:   With  the  coarse flow  off,  set  the fine flow to read each of
     the  four  604 rotameter set points as  noted on the audit form.  Log both
     the  fine  flow rotameter readings and the fine flow in £/min as determined
     by the  last  calibration  for each  of  the 604  rotameter audit set points.

3.    Coarse  flow:   Next,  reset  the  coarse  flow rotameter to  its  normal set
     point  and turn  the fine   flow  completely off.   Replace  the  604 audit
     rotameter with the 603 audit rotameter and  proceed to check each of the 4
     audit  set points.  Log  the 603 rotameter  readings  and the  coarse flow
     rotameter values  and  their corresponding flows  as determined by the last
     calibrations.

Filter Overload Cutout  Switch Check

     Once the  six  month  flow audit  has been  completed,  the filter overload
cutout  switch  setting  should be checked.   Refer  to the instruction for this
check in  Section  5,  Checkout  and Calibration.

     Finally,  return the sampler  to its  normal operation setting.  Void the
filter  set  that was running  on  the day of the  audit and  program the unit  to
                                      76

-------
run on the  next  scheduled sampling day.  Within  one  week after this audit is
performed,  the  impactor  head  assembly  should  be  removed  and  cleaned  in
accordance with the procedure described in Section 8 of this manual.

Six-Month Control Check

     Every  six months control checks  must  be  peformed on at least  5% of the
data  being  collected.  The   checks  must  compare  the  data obtained  from the
computer  output  to the raw  data  recorded on the weekly  logs.   Coarse, fine,
and total mass calculations must be confirmed from raw data.
                                     77

-------
                                    SECTION 8

                                   MAINTENANCE

SCHEDULE FOR ROUTINE MAINTENANCE

     On a  yearly basis  (or  after every  180 24-hour days  of operation),  the
operator should perform the following maintenance:

1.   Replace the Thomas pump diaphrams.

2.   Inspect and replace if  necessary the  shock mounts on  the Thomas pump.

3.   Clean the exhaust filters on the Thomas pump.

4.   Lubricate the lock assembly on the front access door.

5.   Replace the inlet tube connector.

     Every  six  months  (or  after  every  90  24-hour  days  of operation),  the
operator should  perform the following maintenance:

1.   Clean the impactor assembly.

2.   Clean the inlet tube and fractionating  inlet.

3.   Perform a multipoint flow  calibration on the coarse  and  fine  flow  systems
     as described in Section 5  of this  report.
                                      78

-------
MAINTENANCE INSTRUCTIONS ~ YEARLY MAINTENANCE

1.   In performing the suggested yearly maintenance, it is necessary to remove
     the entire  sampler housing  to provide  access  to the  pumps and exhaust
     filters.   Unplug  the sampler  power cord, and remove  the  10 socket head
     cap screws  which attach  the housing to  the  instrument chassis.   Remove
     the  inlet tube,  and carefully  lift  the housing  straight  up  until  it
     clears the instrument.

2.   Diaphram Replacement:

     a.   Remove  the  large  fine  flow accumulator canister located directly
          above the Thomas Pump.

     b.   Disconnect  the polyethylene vacuum  tubing which  is attached to the
          pump with a worm drive  clamp.

     c.   Remove  the  four screws in  each  diaphram head of  the  pump (eight in
          total).

     d.   The  diaphrams  are  visible  once  the pump heads have been  lifted off
          the  pump.

     e.   Replace both diaphragms and reassemble the pump.

 3.   Shock  Mount Inspection:  After  replacing the pump  diaphragms,  thoroughly
     inspect  the  shock  mounts for cracks,  torn  surfaces,  or  other signs of
     fatigue.   Replace  the mounts,  if necessary.

 4.   Exhaust  Muffler  Cleaning:

     a.   After  replacing the pump diaphragms,  run the  pump for five  minutes.
                                      79

-------
     b.    Turn the  pump off and  remove the exhaust mufflers.   Soak  them for
          five  minutes   in  a  solvent  such  as   alcohol.    After   soaking
          thoroughly, dry  the  mufflers  by  blowing them out with  air flow in
          the reverse of  their normal  flow.   Replace the  filters  if  there is
          any indication of clogging.

     c.    Reassemble the system.

5.    Lock Lubrication:   Lubricate the  key-way  on the front  access door with
     Lock Ease,  oil  or some other lubricant designed for locks.

6.    Replacement of Inlet Connector:   Inspect the inlet connector for  signs of
     wear and abrasion.   Replace  the  clear plastic tubing if it is worn or if
     there are any indications  of leaks.  The inlet tube should slide  into the
     plastic connector with  some  resistance;  if the fit is not tight, replace
     the connector tube.

MAINTENANCE INSTRUCTIONS - SEMI-ANNUAL MAINTENANCE

1.    Impactor Assembly Cleaning:

     a.    Remove  the  four screws which attach the  impactor assembly  to the
          instrument deck.

     b.    Disassemble the  impactor by  removing the four screws which  hold the
          impactor  together  and  pulling it  apart,  as  shown  in Figure 26.

     c.    Thoroughly and  gently  clean  all  surfaces of the  impactor and seals
          with  an  ethyl   alcohol  dampened  cloth or  lint-free paper  towel.

     d.    Gently  wipe  the silicon rubber  seals  of  the lower  portion  of the
          filter  seal mechanism  which  remains on the main deck assembly after
          the  impactor  has  been removed  with  an alcohol-dampened  towel  or
          cloth.
                                     80

-------
Figure 26.  Virtual Impactor Disassembled for Cleaning




                                     81

-------
     e.    Reassemble the impactor, making  sure  that the red  rubber  gasket is
          properly  in  place,  before tightening  the screws.  At  all  times be
          careful   never to  place  the  silicon  rubber  filter  seals on  any
          surface  where they may be damaged p_r soiled.
          NOTE:   Unusually  heavy filter loadings resulting  in  values greater
                        o
          than  100  ug/m   will   necessitate  a  more  frequent schedule (e.g.,
          every 45 days).

2.    Inlet Tube and Fractionating Inlet Cleaning:

     a.    Remove the  inlet  assembly from the sampler where it attaches to the
          inlet tube clear plastic connector.

     b.    Disassemble  the  fractionating  inlet and  throughly  wash  it with a
          mild detergent.

     c.    Thoroughly  dry  the  inlet with a  lint free cloth and reassemble the
          unit.
                                      82

-------
                                    SECTION 9

                                 TROUBLESHOOTING

     Suggestions  for  troubleshooting  are  included with  the  checkout  and
calibration  instructions  in  Section  5  of  this  report.    For  additional
troubleshooting instructions, the operator should refer to the manufacturer's
Operating and  Service Instructions  (2).   Problems beyond the  scope  of those
instructions should be referred directly to the manufacturer.
                                     83

-------
                                 REFERENCE LIST

1.    Adapted  from  K.T.   Whitby  and  B.   Cantrell,  "Atmospheric  Aerosols  -
     Characteristics  and   Measurements,"  in   International   Conference  on
     Environmental  Sensing  and  Assessment,   Las  Vegas,  NV,  September 14-17,
     1975, Institute of  Electrical  and Electronics Engineers, New York, 1976.

2.    "Automated  Dichotomous  Particulate Sampling   System:    Operating  and
     Service Instructions,"  Beckman Instruments, Inc., February 1979.

3.    Olin, John  G., "A  New Virtual  Impactor (Dichotomous Sampler)  for Fine
     Particle Air Quality Monitoring," presented at the 71st Annual Meeting of
     the Air Pollution Control Association, Houston, Texas, June 1978.

4.    Loo,  B.W.,  Adachi,  R.S.,   Cork,  C.P.,   Goulding, F.S.,  Jaklevic, J.M.,
     Landis, D.A.,  and  Searles,  W.L.  "A  Second Generation Dichotomous  Sampler
     for  Large  Scale  Monitoring  of  Airborne  Particulate Matter."   Lawrence
     Berkeley Laboratory Report  LBL  -  8725, Berkeley,  CA, January 1979.

5.   "Ambient  Air   Quality  Monitoring,   Data  Reporting,  and  Surveillance
     Provisions,"   U.S.  Environmental  Protection  Agency,  Federal   Register,
     44:27557, May  10, 1979.

6.   Ludwig,  F.L.  Kealoha,  J.H.S.,  and Shelar,  E. ,  "Selecting  Sites for
     Monitoring   Total   Suspended    Particulates,"   EPA    Publication   No.
     EPA-45012-118,  June 1977.
                                      84

-------
                                   APPENDIX I


                              PARTS AND ACCESSORIES


OPERATING ACCESSORIES


1.   Vacuum Gauge:


          1 ea. #V500 V LM #47103

          30" Hg Vacuum Gauge with Screw Check
         Available from:     Boston Cooper
                             Salem Industrial Park
                             Manor Parkway
                             Salem, New Hampshire 03079
                             Tel: (603)893-9181

2.   Pump Muffler:
          2 ea #4450 KI 1/8" Pipe Thread

          Combination muffler/filter
    Available from:     McMaster Carr Supply Co.
                        Post Office Box 4355
                        Chicago, IL 60680
                        Tel: (312) 833-0300

 3.    Inlet  Connector:
           3  feet #PV2620-1  IV  Nominal  ID x 3/16" wall

           Par  Fex Clear  Plastic Tubing
                                      85

-------
    Available from:      Green Rubber Company
                        160 Second Street
                        Cambridge, MA 02142
                        Tel:  (617) 547-7658
4.   Standpipe Extension:

          As  mentioned  in  Section  4,  installation  of  the Sampler  indoors
          requires  an  extension  to  the  inlet  standpipe  and  a  supporting
          sleeve.   Since  the length  and material  for  these accessories will
          depend on the location and type of sampling required, specifications
          are  not  provided  here.   The standpipe  extension  should be  of IV
          nominal  outside diameter  pipe  with  a  1"  nominal  inner diameter.
          Depending on the material selected, pipe and tubing for this purpose
          should   be   readily  available   locally at   heating  or  plumbing
          suppliers.

CALIBRATION ACCESSORIES

1.   Singer Dry  Test  Meter:

          1  ea.   Dry  Test  Meter Model  #DTM-115-3 with  Dial  Temperature and
          Pressure  Gauge

     Available  from:     American  Meter  Division
                        13500 Philmont  Avenue
                        Philadelphia,  PA 19116
2.   Rotameter Assembly (604):

          1 ea.  Model #7630H Rotameter with 604  Tube
          1 ea.  Model #7966  Bench Stand
          1 ea.  h"  diameter  Tantalum Float for  604 Tube
                                      86

-------
    Available from:      Matheson Gas Co.
                        Post Office Box 1147
                        61 Grove Street
                        Gloucester, MA 01930
                        Tel: (617) 283-7700

3.    Rotameter Assembley (603):


          1 ea Model #7630H Rotameter with 603 Tube

          1 ea. Model #7966 Bench Stand


    Available from:      Matheson Gas Co.
                        (See address above)

4.    Inlet Filter for 603 and 604 Rotameters:


          2 ea. #4450 Kl 1/8" Pipe Thread

          Combination muffler/filter


    Available from:      McMaster-Carr Company
                        Post Office Box 4355
                        Chicago,  IL 60680
                        Tel: (312) 833-0300

5.   Inlet Adaptor Assembly (Instructions are given after the specifications):


          6  feet V  ID x V wall  Latex Vacuum Tubing


    Available  from:      Irving  B. Moore
                        30  Rindge Avenue
                        Cambridge, MA 02139
                        Tel: (717) 491-0100

          1  ea. #B-4-FHC-2T  Tapered Female Connector

          2  ea. #8-600-7-4    Female Connector

          3  ea. #B-4-MHC-6T  Tapered Male  Connector

          1  ea. #B-4-HN       Hex Nipple

          1  ea. #B-IRF4-A     Regulating  Valve
                                      87

-------
Available from:     Cambridge Valve and Fitting
                    50 Manning Road
                    Bill erica, MA 01824
                    Tel: (617) 272-8270

      2 feet Schedule 80, 1" Nominal Diameter Type 1 PVC Pipe

      2 ea. Schedule 80, 1" Nominal Diameter Socket Caps

      1 pint "80" PVC Cement
     Available from:     Utility Supply Corporation
                         425 Riverside Avenue
                         Medford, MA
                         Tel: (617) 395-9023

-------
INSTRUCTIONS FOR ASSEMBLY OF INLET CALIBRATION ADAPTOR FOR USE WITH ROTAMETERS

The assembly is shown in Figure 15.

1.    With  PVC  cement,  attach  a  20-inch  length  of Schedule  80  1"  nominal
     diameter pipe  to  a 1" nominal diameter socket cap which has been drilled
     and  tapped to V  NPT.   Remove  all  burrs from the open  end and use the
     second 1" cap as a dust cap (do not glue  it on).

2.    Attach the brass regulating valve's outlet to a V NPT hex nipple.

3.    Attach  the valve and hex  nipple to the  socket cap  on  the tube assembly
     (do not overtighten the brass to PVC connection).

4.    Attach  a  V  NPT tapered  male  connector  to  the   inlet  of  the  brass
     regulating valve.

5.    Use Teflon tape on all pipe-connections.

6.    Attach  the adaptor assembly to  the  604 or 603 rotameter using V  ID and
     V wall latex vacuum  tubing.
                                      89

-------
                                   APPENDIX  II

                      CORRECTION OF ROTAMETER READINGS TO
                         ACTUAL PRESSURE AND TEMPERATURE

      The  flowmeters  described in this  manual  are  manufactured by  Matheson.
 They  come with the manufacturer's calibration  for  air at  70°F  (21.1°C)  and  at
 atmospheric  pressure.   We  recommend  recalibration  of  all  flow  measurement
 devices  to a  traceable reference standard.  The  procedure described below  is
 required  to convert the calibrated flow  readings  to actual  conditions.

      When calibrated rotameters  are  used  at  another pressure  or at  another
 temperature  from  the  calibration conditions,  it is necessary  to  adjust  flow
 readings  for  these changes.   Departures  from calibration  conditions  result  in
 a change  in drag forces on the float  ball.   These changes  are due primarily  to
 changes  in density (pressure  and/or  temperature  affected) and  changes  in air
 viscosity.   Actual flow (operating conditions)  is  the product  of  the  reading
 for  air  times  the density  factor  times  the change of flow due  to  viscosity
 differences.

ACTUAL FLOW =  FLOW READING x DENSITY  FACTOR x %  OF FLOW CHANGE DUE T0 VISC-
                                                           100
DENSITY FACTOR (DF)

             Tl     P?  fc
     DF =  [ -i     -? ]*
             T      P
             '2     Hl

 T!   =    294.3°K (273.2 + 21.1°C) or calibration temperature
 T~   =    Operating Temperature (°K)
 P,   =    Standard  atmospheric pressure  (14.7  psia) or  calibration pressure
           may be expressed in any units of pressure
 Pp   =    Actual atmospheric pressure

                                      90

-------
 VISCOSITY  FACTOR  (VF)

      The  viscosity factor is calculated  by  dividing the dynamic viscosity  in
 centipoise of  air at operating  conditions by the  density of air at  atmospheric
 pressure.   The  viscosity of  air is  independent of pressure  for  the  normal
 range of  surface  pressures.  The  viscosity of air is a  function of  temperature
 as  shown  in Figure A-l where temperature  is  in degree centigrade and  viscosity
 is  in units of centipoise.
                                                                             _o
      Viscosity is divided by density of air  at atmospheric  presure  (1.2 x 10
 grams per cubic  centimeter).   The  simplified formula for the viscosity factor
 includes  these terms times a pressure and temperature correction term.

          VF =  833 (Viscosity)/DF

 Figure A-2 is  then used to determine the percent of flow calculated.   This  is
 determined by  following  the viscosity factor calculated across  to the "float
 curve" and then  up to the horizontal axis to determine the correction factor.

 Sample Calculations

Air Temperature:    -10°C (14°  F)  or 263.2 °K

Pressure:            850 mb, 637.5  mm Hg, or 29" Hg

Density Factor:     [(294.3/263.2) x (29.0/29.92)]^ = 1.04

Viscosity Factor:    833 x 0.01657/1.04 = 13.04

On Figure A-2 using a glass float gives a percentage correction of 104%

ACTUAL FLOW    =    Reading x 1.04 x 1.04
               =    Reading x 1.08
                    or approximately +8%
                                      91

-------
Sample Calculation

     Calibration temperature  Tj = 25°C (77°F) (298.2°K)
     Actual temperature       T2 =  0°C (32°F) (273.2°K)
     Calibration pressure     P., = 790 mm Hg (31.1 inches Hg)
     Actual pressure          P2 = 650 mm Hg (25.59 inches Hg)

Density Factor    =  [(298.2°K/273.2°K) x (650 tm/790m^ = 0.95

Viscosity Factor = (833 x 0.01709)70.95 = 15

 From  Figure A-2  the  correction  factor  is  essentially  100%  of flow  (or no
 change due to viscosity differences).

      ACTUAL FLOW = Reading x .95 x 1.00

                     or approximately a -5%
                                       92

-------
U>
            .0190
            .0180
            .0170
            .0160
                     Figure A-l (Centipoise) of air as a function of temperature  (°C/°F)

-------
§  a   8   s    s
                                                                                     94
0 US GOVERNMENT PRINTING OFFICE 1961  757-064/0314

-------