EPA 600/D-80-019
              Methods: 501.1 and 501.2
Thursday
November 29, 1979
Part 111
Environmental
Protection Agency
Appendix C
Analysis of Trihalomethanes in
[Srinking Water
        U.S. Er;v!"- < ~ -; ~i !-: > "  •
        Region V. ' '
        230 SDUI- L-: .;'-,'.'•.-? s::oel
        Chicago, HiinciCi  60604

-------
68872   Federal Register / Vol. 44. No. 231  /  Thursday,  November 29, 1979 / Rules and Regulations
                        Analysis of
                        Part I: Th« Analysis of TiihalocoethaiM*
                        in Drinking Water by the Purge and Trap
                        Method
                          1. Scope
                          1.1  This method (1) is applicable in
                        the determination of four
                        trihaiomethanes, i.e. chloroform.
                        dichlorobromomethane,
                        dibromochloromethane. andtromoform
                        in finished drinking water, raw source
                        water, or drinking water in any stage of
                        treatment The concentration of these
                        four compounds is totaled to determine
                        total trihaiomethanes (TTHM).
                          1*2  For compounds other than the
                        above-mentioned trihaiomethanes, or
                        for other sample sources, the analyst
                        must demonstrate the usefulness of the
                        method by collecting precision and
                        accuracy data on actual samples as
                        described (2).
                          1.3  Although the actual detection
                        limits are highly dependent upon the gas
                        chromatographic column and detector
                        employed, the method can be used over
                        a concentration range  of approximately
                        0-5 to 1500 micrograms per liter.
                          1.4  Well in excess of 100 different
                        water supplies have been analyzed
                        using this method. Supplementary
                        analyses using gas caromatography
                        mass spectrametry (GC/MS) have
                        shown that there is no evidence of
                        Interference in the determination of
                        trihaiomethanes (3). For this reason, it is
                        nof necessary to analyze the raw source
                        water as is required with the Liquid/
                        Liquid Extraction Method (4).
                          2, Summary
                          Z2  Trihaiomethanes are extracted
                        by an inert gas which is bubbled through
                        the aqueous sample. The
                        trihaiomethanes, along with other
                        organic constituents which exhibit low
                        water solubility and a vapor pressure
                        significantly greater than water, are
                        efficiently transferred  from the aqueous
                        phase to the gaseous phase. These
compounds are swept from the purging
device and are trapped in a short
column containing a suitable sorbent
After a predetermined period of time,
the trapped components are thermally
desorbed and backflushed onto the head
of a gas chromatographic column and
separated under programmed
conditions. Measurement is
accomplished with a halogen specific
detector such as electrolytic
conductivity or microcoulometric
titration.
  2J3   Confirmatory analyses are
performed using dissimilar columns, or
by mass spectrometry (5).
  2.4   Aqueous standards and
unknowns are extracted and analyzed
under identical conditions in order to
compensate for extraction losses.
  ZS   The total analysis time, assuming
the absence of other organohalides. is
approximately 35 minutes  per sample.
  3. Interferences
  3.1   Impurities contained in the purge
gas and organic compounds outgasing
from the plumbing ahead of the trap
usually account for the majority of
contamination problems. The presence
of such Lnteferences are easily
monitored as a part of the  quality
control program. Sample blanks are
normally run between each set of
samples. When a positive
trihalomethane response is noted in the
sample blank, the analyst should
analyze a method blank Method blanks
are run by charging the purging device
with organic-free water and analyzing in
the normal manner.
  If any trihalomethane is noted in the
method blank in excess of 0.4 ^g/l the
analyst should change the  purge gas
source and regenerate the molecular
sieve purge gas filter. Subtracting the
blank values is not recommended. The
use of non-TFE plastic tubing. non-TFE
thread sealants, or flow controllers with
rubber components should be avoided
since such materials generally out-gas
organic compounds which will be
concentrated in the trap during the
purge operation. Such out-gasing
problems are common whenever new
equipment is* put into service; as time
progresses, minor out-gasing problems
generally cure themselves.
  3-2   Several instances of accidental
sample contamination have been noted
and attributed to diffusion of volatile
organics through the septum seal and
into the sample during shipment and
storage. The sample blank is used as  a
monitor for this problem.
  3.3   For compounds that are not
efficiently purged, such as bromoform.
small variations in sample volume,
purge time, purge flow rate, or purge
temperature can affect the analytical

-------
          Federal Register /  Vol.  44. No. 231 / Thursday,  November 29. 1979 /  Rules and Regulations
                                                                      68673
result Therefore, samples and standards
must be analyzed under identical
conditions.
   3.4  Cross-contamination can occur
whenever high-level and low-level
samples are sequentially analyzed. To
reduce this likelihood, the purging
device and sample syringe should be
rinsed twice between samples with
organic-free water. Whenever an
unusually concentrated sample is
encountered, it is highly recommended
that it be followed by a sample blank
analysis to ensure that sample cross
contamination does not occur. For
samples containing large amounts of
water soluble materials, it may be
necessary to wash out the purging
device with a soap solution, rinse with
distilled water, and then dry in a 105*C
oven between analyses.
   3.5  Qualitative misidentifications are
a problem in using gas chramatographic
analysis. Whenever samples whose
qualitative nature is unknown are
analyzed, the following precautionary
measures should be incorporated into
the analysis.
   3.5.1  Perform duplicate analyses
using the two recommended columns
(4.2.1 and 4.Z2) which provide different
retention order and retention times for
the trihalomethanes and other
organohalides.
   3.3-2  Whenever possible, use GC/MS
techniques which provide unequivocal
qualitative identifications (5}.
   4. Apparatus
   4.1  The purge and trap equipment
consists of three separate pieces of
apparatus: the purging device, trap, and
desorber. Construction details for a
purging device and an easily automated
trap-desorber hybrid which has proven
to be exceptionally efficient and
reproducible are shown in Figures 1
through 4 and described in 4.1.1. through
4.1.3. An earlier acceptable version of
the above-mentioned equipment is
described in (lj.
  4.1.1  Purging Device—Construction
details are given in Figure 1 for an all-
glass 5 ml purging device. The glass frit
installed at the base of the sample
chamber allows finely divided gas
bubbles- to pass through the sample
while the sample is restrained above the
frit Gaseous volumes above the sample
are kept to a minimum to eliminate dead
volume effects, yet allowing sufficient
space for most foams to disperse. The
inlet and exit ports are constructed from
heavy-walled VMnch glass tubing so
that leak-free removable connections
can be made using "finger-tight"
compression fittings containing Teflon
ferrules. The removable foam  trap is
used to control samples that foam.
  4.L2  Trapping Device—-The trap
(Figure 2) is a short gas chromatographic
column which at <35" C retards the
flow of the compounds of interest while
venting the purge gas and, depending on
which sorfaent is used, much of the
water vapor. The trap should be
constructed with a low thermal mass so
that it can be heated to ISO* C in less
than 1 minute for efficient desorption.
then rapidly cooled to room temperature
for recycling. Variations in the Crap ID,
wall thickness, sorfaents, sorbent
packing order, and sorfaent mass could
adversely affect the trapping and
desorption efficiencies for compounds
discussed in this text For this reason, it
is important to faithfully reproduce the
trap configurations recommended in
Figure 2. Traps containing Tenax only,
or combinations of Tenax and other
sorbents  are acceptable for this
analysis.
  4.1.3  Desorber assembly—Details for
the desorber are shown in Figures 3, and
4. With the 6-port valve in the  Purge
Sorb position  (Figure 3], the effluent
from the purging device passes through
the trap where the flow rate of the
organics is retarded. The GC carrier gas
also passes through the 6-port valve and
is returned to  the GC With the 6-port
valve in the Purge-Sorb position, the
operation of the GC is in no way
impaired; therefore, routine liquid
injection analyses can be performed
using the gas chromatograph. After the
sample has been  purged, the 3-port
valve is turned to the desorb position
(Figure 4). In this configuration the trap
is coupled in series with the gaa
chromatographic column allowing the
carrier gas to backfiush the trapped
materials into the analytical column.
Just as the valve is actuated, the power
is turned on to the resistance wire
wrapped around the trap. The power is
supplied by an electronic temperature
controller. Using this device, the trap is
rapidly heated to ISO* C and then
maintained at ISO* C with minimal
temperature overshoot The trapped
compounds are released as a "plug" to
the gas chromatograph. Normally,
packed columns with theoretical
efficiencies near 500 plates/foot under
programmed temperature conditions can
accept such desorb injections without
altering peak geometry. Substituting a
non-controlled power supply, such as a
manually-operated variable transformer,
will provide nonreproductible  retention
times and poor quantitative data unless
Injection Procedure (8.3.2] is used.
  4.1.4  Several Purge and Trap Devices
are now commercially available. It is
recommended that the following be
taken into consideration if a unit is to be
purchased:
  a. Be sure that the unit is completely
compatible with the gas chromatograph
to be used for the analysis.
  b. Use a S-ml purging device similar to
that shown in Figure 1.
  c. Be sure the Tenax portion of the
trap meets or exceeds the dimensions
shown in Figure 2.
  d. With the exception of sample
introduction, select a unit that has as
many of the purge trap functions
automated as possible.
  4.2   Gas chromatograph—The
chromatograph must be temperature
programmable and equipped with a
halide specific detector.
  4.2.1  Column I is an unusually
efficient column which provides
outstanding separations for a wide
variety of organic compounds. Because
of its ability to resolve trihalomethanes
from other organochlorine compounds,
column I should be used as the primary
analytical column (see Table 1 for
retention data using'this column}.
  4.2.1.1  Column I parameters:
Dimensions—3 feet long x 0.1 inch ID
stainless steel or glass tubing. Packing—
1% SP-1000 on Carfaopack-B (60/80)
mesh. Carrier Gas—helium at 40 ml/
minute. Temperature program sequence:
45* C isothermal for 3 minutes, program
at 8* C/minute to 220* C then hold for 15
minutes or until all compounds have
eluted.
  Note^-It has been found that during
handling, picking, and programming, active
sites an exposed on the Carbopack-8
packing. This results in tailing peak geometry
and poor resolution of many constituents. To
correct this, pack the first 5 cm of the column
with 3* SIMOOO on Oiromosorb-W 60/80
followed by the Carbopack-3 packing.
Condition the preeolumn and the Carbopack
columns with  carrier gas flow at 220* C
overnight Pneumatic shocks and rough
treatment of packed columns will cause
excessive fracturing of the Carbopack. If
pressure in excess of 60 psi is required to
obtain 40 ml/minute earner flow, then the
column should be repacked.
  4.2.1.2  Acceptable column  equivalent
to Column I; Dimensions—3 feet
long x 0.1 inch ID stainless steel or glass
tubing. Packing—0.2% Carbowax 1500
on Carbopack-C (80/100) mesh. Carrier
Gas—helium at 40 ml/minute.
Temperature program sequence—30* C
isothermal for 3 minutes, program at 8*
C /minute to 160* C then hold for 2
minutes or until all compounds have
eluted.
  Note.—It has been found that during
handling, packing, and programming, active
sites are exposed on the Carbopack-C
packing. This results in poor resolution of
constituents and poor peak geometry. To
correct this, place a 1 ft. 0.125 in. CD x 0.1 in.

-------
 68674   Federal Register / Vol. 44. No. 231 / Thursday.  November 29. 1979  /  Rules  and Regulations
 ID stainless steel column packed with 3*
 Carbowax 1500 on Chromosorb-W 80/80
 mesh in series before the Carbopack-C
 column. Condition the precolumn and Che
 Carbopack columns with carrier gas flow at
 190* C overnight The two columns nay b«
 retained in series for routine analyses.
 Trihaiomethane-retention times are listed in
 Table 1.
   4.22  Column H provides unique
 organohalide-trihalomethane
 separations when compared to those
 obtained from Column I (see Figures 5
 and 6). However, since the resolution
 between various compounds is generally
 not as good as  those with Column I it is
 recommended that Column II be used as
 a qualitative confirmatory column for
 unknown samples when GC/MS
 confirmation is not possible.
   4.Z2.1  Column II parameters:
 Dimensions—8 feet long x 0.1 inch ID
 stainless steel or glass. Packing—n-
 octane on Porisil-C (100/120 mesh).
 Carrier Gas—helium at 40 cc/minute.
 Temperature program sequence—30* C
 isothermal for 3 minutes, program at 9*/
 minute to 170' C then hold for 4 minutes
 or until all compounds have eluted.
 Trihalomethane retention times are
 listed in Table 1.
   5.8  Organic-free water is defined as
 water free of interference when
 employed in the purge and trap analysis.
   5.8.1  Organic-free water is generated
 by passing tap water through a carbon
 filter bed containing about l Ib. of
 activated carbon. Change the activated
 carbon bed whenever the concentration
 of any trihalomethane exceeds 0.4 pg/L
   5^2  A Miilipore Super-Q Water
 System or its equivalent may be used to
 generate organic-free water.
   SUU  Organic-free water may also be
 prepared by boiling water for 15
 minutes. Subsequently, while
 maintaining the temperature at 90* C.
 bubble a contaminant-free inert gas
 through the water for one hour. While
 still hot transfer the water to a narrow-
 mourn screw-cap bottle with a Teflon
 seal
  5.8.4  Test organic free water each
 day it is used by analyzing according to
 Section 3.
  5.9   Standards.*
  3.9.1  Bromofonn—96%—available
 from Aldrich Chemical Company.
  5.9.2  Bromodichloromethane 97%—
 available from Aldrich Chemical
 Company.
  5.9.3  Chiorodibromomethane—
 available from Columbia Chemical Inc.
 Columbia. S.C
  5.9.4  Chloroform—9995—available
from Aldrich Chemical Company.
  'Ait precautionary nuuun. all standards must
b« checked for punty by boiling point
diltrnunaiioni or GC/MS a»*yi (i).
   5.10  Standard Stock Solutions
   3.10.1  Place about 9.3 ml of methyi
 alcohol into a ground glass stoppered 10
 ml volumetric flask.
   5.10.2  Allow the flask to stand
 unstoppered about 10 minutes or until
 all alcohol wetted surfaces have dried.
   5.10.3  Weigh the flask to the nearest
 0.1 mg.
   5.10.4  Using a 100 pi syringe.
 immediately add 2 drops of the
 reference standard to the flask, then
 reweigh. Be sure that the 2 drops fall
 directly into the alcohol without
 contacting the neck of the flask.
   5.10.3   Dilute to volume, stopper, then
 mix by inverting the flask several times*
   5.10.6   Transfer the solution to a
 dated and labeled 15 ml screw cap
 bottle with a Teflon cap liner.
   Note—Because of the toxicity of
 trihalomethanca. it is necessary to prepare
 primary dilutions in a hood. It is further
 recommended that a NIOSH/MESA
 approved toxic gas respirator be used when
 the analyst handles high concentrations of
 such materials.
   5.10.7   Calculate  the concentration in
 micrograms per microliter from the  net
 gain in weight
   5.10.8   Store the solution at 4* C.
  Note.—All standard solutions prepared in
 methyl alcohol are stable up to 4 weeks when
 stored under these conditions. They should
 be discarded after that time has elapsed.
   5.11  Aqueous Calibration Standard
 Precautions.
   5.11.1   In order to prepare accurate
 aqueous standard solutions, the
 following precautions must be observed.
   a. Do not inject more than 20 pi of
 alcoholic standards  into 100 ml of
 organic-free water.
  b. Use of 23 pi Hamilton 702N
 nu'crosyringe or equivalent (Variations
 in needle geometry will adversely affect
 the ability to deliver reproducible
 volumes of methanolic standards into
 water.)
  c. Rapidly inject the alcoholic
 standard into the expanded area of the
 filled volumetric flask. Remove the
 needle as fast as possible after injection.
  d. Mix aqueous standards by inverting
 the flask three times only.
  e. Discard the contents contained in
 the neck of the flask. Fill the sample
 syringe from the standard solution
 contained in the expanded area of the
 flask as directed in Section 3.5.
  f. Never use pipets to dilute or transfer
samples or aqueous  standards.
  g. Aqueous standards when stored
with a headspace are not stable and
should be discarded after one hour.
  h. Aqueous standards can be stored
according to Sections 8.4 and 3.8.
  5.11.2  Prepare, from the standard
stock solutions, secondary dilution
 mixtures in methyl alcohol so that a 20
 pi injection into 100 ml or organic-free
 water will generate a calibration
 standard which produces a response
 close (±10%) to that of the sample (See
 9.1).
   5.11.3  Purge and analyze the
 aqueous calibration standards in the
 same manner as the samples.
   5.11.4  Other calibration procedures
 (3) which require the delivery of less
 than 20 pi of a methanolic standard into
 a 5.0 ml volume of water already
 contained in the sample syringe are
 acceptable only if the aiethanoiic
 standard is delivered by  the solvent
 flush technique (6).
   5.12  Quality Check Standard (2.0 pg/
 1)
   5.12.1  Prom the standard stock
 solutions, prepare a secondary dilution
 in methyi alcohol containing 10 ng/pl of
 each trihalomethane (See Section 5.10.3
 Note).
   5.12.2  Daily, inject 20.0 pi of this
 mixture into 100.0 ml of organic-free
 water ana analyze according to Section
 3.
  8. Sample Collection and Handling
  9.1.  The sample containers should
 have a total volume of at least 25 ml.
  8.1.1  Narrow mouth screw cap
 bottles with the TFE fluorocarbon face
 siiicone sepata cap liners are strongly
 recommended.
  8.2   Sample Bottle Preparation
  6.2.1  Wash all sample bottles and
 TFE seals in detergent Rinse with tap
 water and finally with distilled water.
  8.2^  Allow the bottles and seals to
 air dry at room temperature, then place
 in  a 105*  C oven for one hour, then allow
 to  cool in a area known to be free of
 organics.
 Note.—Do not heat the TFE seals for
 extended period of time (>l hour) because
 th« silicons Uy«r slowly degrades at 105* C.
  8.2.3  When cool, seal  the bottles
 using the TFE seals that will be used for
 sealing the samples.
  3.3   Sample Stabilization—A
 chemical reducing agent (Section 5.5) is
 added to  the sample in order to arrest
 the formation of trihalo-methanes after
 sample collection (3, 7). Do not add the
reducing agent to samples when data on
maximum trihalomethane formation is
 desired. If chemical stabilization is
 employed, the reagent is also added to
 the blanks. The chemical agent (2.5 to 3
 mg/40 ml) is added  to  the empty sample
 bottles just prior to shipping to the
 sampling site.
  3.4   Sample Collection
  3.4.1  Collect all samples in duplicate.
  8.4.2  Fill the sample bottles in such a
 manner that no  air bubbles pass through
 the sample as the bottle, is filled.

-------
         Federal  Register / Vol. 44. No. 231  /  Thursday,  November 29, 1979  /  Rules and Regulations   68675
  6.4.3  Seal the bottles so that no air
bubbles are entrapped in it.
  6.4.4  Maintain the hermetic seal on
the sample bottle until analysis.
  6.4.5  Sampling from a water tap.
  8.4.S.1  Turn on water and allow the
system to flush until the temperature of
the water has stabilized. Adjust the flow
to about 500 mi/minute and collect
duplicate samples from the flowing
stream.
  6.4.6  Sampling from an open body of
water.
  8.4.6.1  Fill a 1-quart wide-mouth
battle with sample from a representative
area. Carefully fill duplicate sample
bottles from the 1-quart bottle as noted
in 6.4-2.
  6.4.7  If a chemical reducing agent
has been added to the sample bottles,
fill with sample just to overflowing, seal
the bottle, and shake vigorously for 1
minute.
  9.4.S  Sealing practice for septum seal
screw cap bottles.
  6.4.8.1  Open the bottle and 511 to
overflowing, place on a level surface,
position the TFE side of the septum seal
upon the convex sample meniscus and
seal the bottle by screwing the cap on
tightly.
  8-4A2  Invert  the sample and lightly
tap the cap on a solid surface. The
absence of entrapped air indicates a
successful seal. If bubbles are present
opes the bottle, add a few additional
drops of sample and reseal toe bottfe as
above.
  6.4£  Blanks.
  6.4.9.1  Prepare blanks in duplicate at
the laboratory by filling and sealing
sample bottles with organic-free water
just prior to shipping the sample bottles
to the sampling site.
  6.4.9.2  If the sample is to be
stabilized, add an identical amount of
stabilization reagent to the blanks.
  6.4.9.3  Ship the blanks to and from
the sampling site along with the sample
bottles.
  8.iS.4 Store the blanks  and the
samples collected at a given site (sample
set) together. A sample set is defined as
ail the samples collected at a given site
(Le.. at a water treatment plant, the
duplicate raw source waters, the
duplicate finished waters and the
duplicate blank samples comprise the
sample set).
  0.5  When samples have been
collected according to Section 8. no
measurable loss of thhalomethanes has
been detected over extended periods of
storage time (3). It is recommended that
all samples be analyzed within 14 days
of collection.
  7. Conditioning Traps
  7.1  Condition newly packed traps
overnight at ISO* C with an inert gas
flow of at least 20 ml/min.
  7.1.1  Vent the trap effluent to the
room, not to the analytical column.
  7.2  Prior to daily use. condition traps
10 minutes while backflushing at 180* C
It may be beneficial  to routinely
condition traps overnight while
backflushing at 180*  C
  7.Z1  The trap may be vented to the
analytical column; however, after
conditioning, the column must be
programmed prior to use.
  8. Extraction and Analysis
  3.1  Adjust the purge gas (nitrogen or
helium) flow rate to 40 ml/min.
  8.2  Attach the trap inlet to the
purging device. Turn the valve to the
purge-sorb position (Figure 3).
  3.3  Open the syringe valve located
on the purging device sample
introduction needle.
  8.4  Remove the plungers from two S
mi syringes and attach a dosed syringe
valve to each.
  8.5  Open the sample bottle and
carefully pour the sample into one of the
syringe barrels until it overflows.
Replace the syringe plunger and
compress the sample. Open the syringe
valve and vent any residual air while
adjusting the sample volume to 5.0 ml.
Close the valve.
  8.6  Fill the second syringe in an
identical manner from the same sample
bottle. This second syringe is reserved
for a duplicate analysis, if necessary
(See Sections 9.3 and 9.4).  -
  3.7  Attach the syringe-valve
assembly to the syringe valve on the
purging device.
  8.3  Open the syringe valve and inject
the sample into the purging chamber.
Close both valves. Purge the sample for
ll.0i.05 minutes.
  3.9  After the 11-minute  purge time.
attach the trap to the chromatograph
(turn the valve to the desorb position)
and introduce the trapped materials to
the GC column by rapidly heating the
trap to 180'C while backflushing the trap
with an inert gas between 20 and 60 ml/
min for 4 minutes.
  8.9.1  If the trap can be rapidly
heated to 180'C and maintained at this
temperature, the GC analysis can begin
as the sample is desorbed. i.e^ the
column is at die initial 45*C operating
temperature. The equipment described
in Figure 4 will perform accordingly.
  3.9.2  With other  types of equipment
(see Section 4.1.4 and Reference 1)
where die trap is not rapidly heated or is
not heated in a reproducible manner, it
may be necessary to transfer the
contents of the trap  into the analytical
column at <30*C where it  is once again
trapped. Once the transfer is complete (4
minutes), the column is rapidly heated to
the initial operating temperature for
analysts.
  8.9.3  If injection procedure 8.9.1 is
used and the early eluting peaks in the
resulting chromatogram have poor
geometry or variable retention times,
then Section 8.9.2 should be used.
  8.10  After me extracted sample is
introduced into the gas chromatograph.
empty the gas purging device using the
sample introduction syringe, followed
by two 5-ml flushes of organic-free
water. When the purging device is
emptied, leave the syringe valve open
allowing the purge gas to vent through
the sample introduction needle.
  3.11  Analyze each sample and
sample blank from the sample set in an
identical manner (see Section 8.4.9.4) on
the same day.
  8.12  Prepare calibration standards
from the standard stock solutions
(Section 5.10) in organic-free water that
are close to the unknown in
trihalomethane composition and
concentration (Section 9.1). The
concentrations should be such that only
20 pi or less of the secondary dilution
need be added to 100 ml of organic-free
water to produce a standard at the same
level as the unknown.
  8.13  As an alternative to Section
8.12. prepare a calibration curve for
each trihalomethane containing at least
3 points, two of which must bracket the
unknown.
  9. Analytical Quality Control
  9.1 Analyze the 2 fig/1 check sample
daily before any samples are analyzed.
Instrument status checks and lower limit
of detection estimations based upon
response factor calculations at five
times the noise level are obtained from
these data. In addition, response factor
data obtained from the 2 /ig/1 check
standard can be used to estimate the
concentration of the unknowns. From
this information, the appropriate
standard dilutions can be determined.
  9.2 Analyze the sample blank to
monitor for potential interferences as
described in Sections 3.1. 3.2. and 3.4.
  9.3 Spiked Samples
  9.3.1  For laboratories analyzing more
than 10 samples a day, each 10th sample
should be a laboratory generated spike
which closely duplicates the average
finished drinking water in
trihalomethane composition and
concentration. Prepare the spiked
sample in organic-free water as
described in Section 3.11.
   9.3.2  For laboratories analyzing less
than 10 samples daily, each time the
analysis is performed, analyze at least 1
laboratory generated spike sample
which closely  duplicates the average
finished drinking water in

-------
 68876   Federal Register  /  Vol. 44. No. 231  /  Thursday. November 29. 1979 / Rules, and Regulations
 trihalomethane composition and
 concentration. Preoare the spiked
 sample in organic-free water as
 described in Section 5.11.
   9.4  Randomly select and analyze
 10% of all samples in duplicate.
   9.4.1  Analyze all samples in
 duplicate which appear to deviate more
 than 30% from any established norm.
   9.5  Maintain an up-to-date log on the
 accuracy and precision data collected in
 Sections 9.3 and 9.4. If results are
 significantly different than those cited in
 Section 11.1. the analyst should check
 out the entire analyses scheme to
 determine why the laboratory's
 precision and accuracy limits are
 greater.
   9.9  Quarterly, spike an EMSL-
 Cincinnati tnhalomethane quality
 control sample into organic-free water
 and analyze.
   9.6.1 The results  of the EMSL
 trihalomethane quality control  sample
 should agree within  20% of the  true
 value for  each tnhalomethane.  If they do
 not then the analyst must check each
 step in the standard generation
 procedure to solve the problem (Section
 3.9. 5.10. and 5.11).
   9.7  Maintain a record of the
 retention  times for each trihalomethane
 using data gathered  from spiked
 samples and standards.
   9.7.1  Daily calculate the average
 retention  time for each  trihalomethane
 and the variance encountered for the
 analyses.
   9.7.2  If individual trihalome thane
 retention  time varies by more than 10%
 over an eight hour period or does not fall
 with 10% of an established norm, the
 system is  "out of control." The source of
 retention  data variation must be
 corrected before acceptable'data can  be
 generated.
   10. Calculations
   10.1  Locate each trihalomethane in
 the sample chromatogram by comparing
 the retention time of the suspect peak to
 the data gathered in  9.7,1.  The retention
 time of the suspect peak must fall within
 the limits  established in 9.7.1 for single
column identification.
   10.2  Calculate the concentration of
the samples by comparing the peak
height or peak areas of  the samples to
the standard peak height (8112). Round
off the data to the nearest jig/1 or two
significant figures.
        MM ntiqm lamsia

        Mai n*qm staneare
[cone. sta. »o,/i)
                 10.3  Report the results obtained from
               the lower limit of detection estimates
               along with the data for the samples.
                 10.4  Calculate the total
               trihalomethane concentration (TTHM)
               by summing the 4 individual
               trihalomethane concentrations in ng/1.
               TTHM (>ig/l)-(Conc, CHCl,)-r(Conc.
               CHBrCl«)-i-(Conc. CHBrjCl]-t-(Conc.
               CHBr).'
                 10.5  Calculate the limit of detection
               (LOO] for each trihalomethane not
               detected using the following criteria:
                          /  AXATT \
                LOG (n4/n-  (	1 a v#n
                          \  SxATT /
               where 8—peak  height (nun) of 2 jj.g/1 quality
                   check standard
               A-3 times the noise level in (mm) at the
                   exact retention time of the
                   tnhalomethane or the baseline
                   displacement in (mm) from the
                   theoretical zero at the exact retention
                   time of the trihalomethane.
               ATT—Attenuation factor
                 11. Accuracy and Precision
                 11.1   One liter of organic-free water
               was spiked with the trihalomethanes
               and used to fill septum seal vials which
               were stored under ambient conditions.
               The spiked samples, were randomly
               analyzed overa 2-week period of time.
               The single laboratory data listed in
               Table a reflect the errors due to the
               analytical procedure and storage.
              'References
                 1. Bellar. T. A_ f. ]. lichtenberg.
               Determining Volatile Organic* at the
               Microgram per Litre Levels by Gas
               Chromatography. Journal AWWA.. id. 739
               (December 1974).
                 Z. "Handbook for Analytical Quality
               Control in Water and Wastewater
               Laboratories." Analytical Quality Control
               Laboratory. National Environmental
               Research Center. Cincinnati. Ohio. June 1972.
                 3. Brass, H. J.. et ai., "National Organic
               Monitoring Survey: Sampling and Purgeable
               Organic Compounds. Drinking Water Quality
               Through Source Protection." R. B. Pojasak.
               Editor. Ann Arbor Science, p. 398.1977
                 4. 'The Analysis of Trihalomethanes in
               Finished Water by the Liquid/Liquid
               Extraction Method. Method 501.2"
               Environmental Monitoring and Support
               Labortory. Environmental Research Center.
               Cincinnati. Ohio. 43288. May 15.1979.
                 5. Budde. W. L and f. W. Eicheiberger.
               "Organics Analysis Using Cas
               Chromatography-Mass Spectrometry." Ann
               Arbor Science. Ann Arbor. Michigan. 1979.
                 9. White. L D. et at, "Convenient
               Optimized Method for the Analysis of
               Selected Solvent Vapors in the Industrial
               Atmosphere." AIHA Journal VoL 31. p. 223.
               1970.
                 7  Kopfler. F. C. at aL "GC/MS
               Determination of Volatile! for the National
               Organics Reconnaissance Survey (NORS)  or
Drinking Water. Identification and Analysis
of Organic Pollutants in Water." L. H. Keith.
Editor. Ann Arbor Science, p. 37.1976.
   Tabto l—3«r*rmon Oau for Tnttaiorrutntnei
                    flcumon am* mrnuiM
                       AccaeueM
                       Altamaow
                Catumn I    to     Column n
                i%sotOOO caturmi   n-octan*
               CaroopacK 3  04%
                       CanMwa>
                       CareeeacK
araimdKnMranwman*
CMcfOflUwmonwtnaiw
  (Dibfomocmcfomamanet
                    10.7
                    13.7
                    185
                    192
 32
10.3
:32
•37
166
•92
         ng(» laooratory Accuracy and Prscoton
SO** NufllMr
rt/l lampia*
12 ..
110
119.0
1 6 .
160
t«00
20
20.0
196.0 '.

13 .
230
23i a ...
12
3
11
12
3
11
.*...,
GntOf
12
3
11

1Z
a
n
Pfacuion Accuracy
Maan landard pvctni
>t9/l daviaoon ricovary
12
11
tOS
BQcNOfoniafftana
13
!3
t4S.
19
19
'83.
Sreinehxni
13
23
233
014
0.16
79
OOS
0.39
10.2
0.09
0.70
103

016
1 38
163
100
92
aa
9*
94
31
93
95
9«

too
'00
37
BHJJNG COOI (360-0t-M

-------
    Federal Register / Vol. 44. No. 231 / Thursday. November 29. 1979 / Rules and Regulations   68677
            OPTIONAL
            FOAM TRAP
IN. O.D. EXIT
  EXIT 1/4
  IN. O.D.

— 14MM. O.D.

INLET 1/4
IN. O.D.
       10MM. GLASS FRIT
       MEDIUM POROSITY
                                     SAMPLE INLET

                                       2-V/AY  SYRINGE VALVE
                                      17CM. 20 GAUGE SYRINGE NEEDLE
                                     6MM.O.D.  RUBBER SEPTUM
                                         10MM.  O.D.
                                           INLET
                                           1/4 IN. O.D.
                           T/16 IN. O.D.
                           STAINLESS STEEL
                                                         13X MOLECULAR
                                                         SIEVE PURGE
                                                         GAS FILTER
                                                            PURGE GAS
                                                            FLOW CONTROL
                FIGURE 1;   PURGING  DEVICE

-------
66678
Federal Register / Vol. 44. No. 231 / Thursday. November 29. 1979 /  Rules and Regulations
O

~
O
u
                                  \J    \J   \J   \J








at
a
u
O
ae

2
e.


5
<
*»
o





fe
<

S
^
c
3
1



Q {v^.*;---' T. .• «• - •; v.'r

s
s
M
S
o
$
V*
<
*tf


0 r^$$$^^/^'?:\;-::

*
£ s
w
S
i --
* -o
•• a <
* y^ Kc;~i^H; ys<7^:^?j^

s
w
«%

at
2
j^

5 1
. w
5
_ o
• <£
o s

- e


H)

- S
5 s
<• •*
5
Q
J *
Kf
** O
s
<
«








•»
s
s
t.
«




                                                                                  O

-------
         CARRIER GAS FLOW CONTROL




PRESSURE REGULATOR
                                    LIQUID INJECTION PORTS
                                                           COLUMN OVEN
PURGE GAS     .

FlOW CONTROL  X
I3X MOLECULAR

SIEVE FILTER
                                        '  |~] fl n n	L -,	CONFIRMATORY COLUMN

                                        I U  LJ U  U   I X Tn nnFf ion
                                        umnr-t
                                                        TO DETECTOR



                                                           ANALYTICAL COLUMN
                                       OPTIONAL 4-PORT COLUMN

                                       SELECTION VALVE


                                         TRAP INLET (fENAX END)


                               6-PORT VALVE I RESISTANCE WIRE
                                                                     HEATER CONTROL
                                                           NOTE: ALL LINES BETWEEN

                                                                TRAP AND  GC

                                                                SHOULD  DE HEATED

                                                                TO 00°C
                                               PURGING DEVICE
             FIGURE 4 PURGE-TRAP SYSTEM (DESORB MODE)
                                                                                 I
4
M
ST
                                                                                    c
                                                                                    O-

                                                                                    01
                                                                                 o
                                                                                 <
                                                                                 n>
                                                                                 a
                                                                                 a-
                                                                                 E.
                                                                                 o>
                                                                                 u

                                                                                 PI
                                                                                 a
                                                                                 a.

-------
68680     Federal Register / Vol. 44. No. 231 /  Thursday. November 29. 1979 / Rales and Regulations
                                                                                 Q
                                                                                 o
                                                                                 S3
                                                                                 si
                                                                                 O
                                                                                 tn
                                                                                   i
                                                                                 LU
                                                                                 O
                                                                                 LU
                                                                                 UJ
                                                                                 O
                                                                                 esi
                                                                                 uu
                                                                                 O

-------
Federal Register / Vol. 44. Mo. 231 / Thursday, November 29. 1979 / Rules and Regulations    68681

-------
68682
Federal Register / Vol. 44. No. 231 / Thursday. November 29. 1979 / Rules and Regulations
                              : 11 - JNT-UOIOHOHI  m
                                INTMUWOIO1IC T MUO1M3
I

I
z
                                                                               Q
                                                                               _j
                                                                               <


                                                                               O
                                                                                o
                                                                                es,
                                                                                O
                                                                                u.
                                                                                o
                                                                      o
                                                                      O
                                                                      *~
                                                                      <
                                                                      s
                                                                      o
                                                                                2
                                                                                O
                                                                                     §
                                                                                     W
                                                                                     i
                                                                                     i

-------
Federal  Register / Vol. 44.  No. 231 / Thursday. November  29. 1979 / Rules and Regulations   68683
Part Q: Analysis of Trihalomethanes in
Drinking Water by Liquid/Liquid
Extraction
  1. Scope.
  1.1  This method (1.2) is applicable
only to the determination of four
trihalomethanes. i.e- chloroform,
bromodichlorome thane.
chlorodibromomethane, and bromoform
in finished drinking water, drinking
water during intermediate stages of
treatment and the raw source water.
  1.2  For compounds other than the
above-mentioned trihalomethanes. or
for other sample sources, the analyst
must demonstrate  the usefulness of the
method by collecting precision and
accuracy data on actual samples as
described in (3) and provide qualitative
confirmation of results by Gas
Chroma tography/Mass Spectrometry
(GC/MS) (4).
  1.3  Qualitative analyses using GC/
MS or the purge and"trap method (5)
must be performed to characterize each
raw source water if peaks appear as
interferences in the raw source analysis.
  1.4  The method has been shown to
be useful for the trihalomethanes over a
concentration range from approximately
0.5 to. 200 fig/L Actual detection limits
are highly dependent upon the.
characteristics of the gas
chromatographic system used.
  2. Summary
  2.1  Ten milliliters of sample are
extracted one time with 2 ml of solvent
Three fil of the extract are then injected
into a gas chromatograph equipped with
a linearized electron capture detector
for separation and analysis.
  2-2  The extraction and analysis time
is 10 to 50 minutes per sample
depending upon the analytical
conditions chosen. [See Table 1 and
Figures 1, 2. and 3.)
  2.3  Confirmatory evidence is
obtained using dissimilar columns and
temperature programming. When
component concentrations are
sufficiently high (>50 fig/1), halogen
specific detectors may be employed for
improved specificity.
  2.4  Unequivocal confirmatory
analyses at high levels (>50 fig/1) can
be performed using GC/MS in place of
the electron capture detector.  At levels
below SO fig/L unequivocal confirmation
can only be performed by the purge and
trap technique using GC/MS [4. 5).
  2.5  Standards dosed into organic
free water and the samples are
extracted and analyzed in an  identical
manner in order to compensate for
possible extraction losses.
  2.8  The concentration of each
trihalomethane is summed and reported
as total trihalomethanes in fig/L
                                3. Interferences
                                3.1  Impurities contained in the
                              extracting solvent usually account for
                              the majority of the analytical problems.
                              Solvent blanks should be analyzed
                              before a new bottle of solvent is used to
                              extract samples. Indirect daily checks
                              on the extracting solvent are obtained
                              by monitoring the sample blanks (9.4.10).
                              Whenever an interference is noted in
                              the sample blank, the analyst should
                              reanalyze the extracting solvent The
                              extraction solvent should be discarded
                              whenever a high level (>10 fig/1) of
                              interfering compounds are traced to it
                              Low level interferences generally can be
                              removed by distillation or column
                              chromatography (6): however, it is
                              generally more economical to obtain a
                              new source of solvent or select one of
                              the approved alternative solvents listed
                              in Section 5.1. Interference free solvent
                              is defined as a solvent containing less
                              than 0.4 fig/1 individual trihalomethane
                              interference. Protect interference-free
                              solvents by storing in a non-laboratory
                              area known to be free of organochlorine
                              solvents. Subtracting blank  values is not
                              recommended.
                                3.2  Several instances of  accidental
                              sample contamination have been
                              attributed to diffusion of volatile
                              organics through the septum seal on the
                              sample bottle during shipment and
                              storage. The sample blank (8.4.10) is
                              used to monitor for this problem.
                                3.3  This liquid/liquid extraction
                              technique efficiently extracts a wide
                              boiling range of non-polar organic
                              compounds and. in addition, extracts the
                              polar organic components of the sample
                              with varying_fifSciencie3. In order to
                              perform the trihalomethane  analysis as
                              rapidly as possible with sensitivities in
                              the low fig/1 range, it is necessary to use
                              the semi-specific electron capture
                              detector and chromatographic columns
                              which have relatively poor resolving
                              power. Because of these concessions,
                              the probability of experiencing
                              chromatographic interferences is high.
                              Trihalomethanes are primarily products
                              of the chlorination process and
                              generally do not appear in the raw
                              source water. The absence of peaks in
                              the raw source water analysis with
                              retention times similar to the
                              trihalomethanes is generally adequate
                              evidence of an interference-free finished
                              drinking water analysis. Because of
                              these possible interferences, in addition
                              to each finished drinking water analysis,
                              a representative raw source water (6.4.5)
                              must be analyzed. When potential
                              interferences are noted in the raw
                              source water analysis, the alternate
                              chromatographic columns must be used
                              to reanalyze the sample set If
interferences are still noted, qualitative
identifications should be performed
according to Sections Z2 and 2.4. If the
peaks are confirmed to be other than
trihalomethanes and add significantly to
the total trihalomethane value in the
finished drinking water analysis, then
the sample set must be analyzed by the
purge and trap method (5).

4. Apparatus
  4.1  Extraction vessel—A15 ml total
volume glass vessel with a Teflon lined
screw-cap is required to efficiently
extract the samples.
  4.1.1  For samples that do not form
emulsions 10 ml screw-cap flasks with a
Teflon faced septum (total volume is ml)
are recommended. Flasks and caps—
Pierce—#13310 or equivalent Septa—
Teflon silicons—Pierce #12718 or
equivalent.
  4.1.2  For samples that form
emulsions (turbid source  water) 15 ml
screw cap centrifuge tubes with a Teflon
cap liner are recommended. Centrifuge
tube—Coming 8062-15 or equivalent
  4.2  Sampling containers—40 ml
screw cap sealed with Teflon faced
silicone septa. Vials and  caps—Pierce
#13075 or equivalent Septa—Pierce
#12722 or equivalent
  4.3  Micro syringes—10.100 fil.
  4.4  Micro syringe—25 fil with a 2-
incn by 0.006-inch needle—Hamilton
702N or equivalent
  4.5  Syringes—10 ml glass
hypodermic with luerlok  tip (2 each).
  4.8  Syringe valve—2-way with luer
ends (2 each)—Hamilton =86570—1FM1
or equivalent
  4.7  Pipette—2.0 ml transfer.
  4.8  Glass stoppered volumetric
flasks—10 and 100 ml
  4.9  Gas chromatograph with
linearized electron capture detector.
(Recommended option—temperature
programmable. See Section 4.12.)
  4.10  Column A—1 mm ID x 2m long
glass packed with 3% SP-1000 on
Supelcoport (100/120 mesh) operated at
50'C with 80 ml/min flow. (See Figure 1
for a sample chromatogram and Table 1
for retention data.)
  4.11  Column B—2 mm ID x 2m long
glass packed with 10% squaiane on
Chromosorb WAW (80/100 mesh)
operated at 67*C with 25 ml/min flow.
This column is recommended as the
primary analytical column.
Trichloroethylene, a common raw
source water contaminate, coelutes with
bromodichloromethane. (See Figure 2 for
a sample chromatogram  and Table 1  for
retention data.)
  4.12  Column C—2 mm ID x'3m long
glass packed with 6% OV-n/4% SP-
2100 on Supelcoport (100/120 mesh)
temperature program 45* C for 12

-------
68S84   Federal Register / Vol. 44. No.  231 / Thursday.  November 29. 1979  / Rules  and  Regulations
 minutes, then program at I'/minute to
 70'C with a 25 ml/sain flow. (See Figure
 3 for a sample chromatogram and Table
 I for retention data.)
   4.13  Standard storage containers—15
 ml amber screw-cap septum bottles with
 Teflon faced silicone septa. Bottles and
 capo—Pierce *19830 or equivalent
 Septa—Pierce #12716 or equivalent.

 5. Reagents
   5.1  Extraction solvent—(See 3.1).
 Recommended—Pentane*. Alternative—
 hexane. methylcydohexane or &Z4-
 trimethylpentane.
   5.2  Methyl alcohol—ACS Reagent
 Grade.
   5.3  Free and combined chlorine
 reducing agents—Sodium thiosulfata
 ACS Reagent Grade—sodium sulfite
 ACS Reagent Grade.
  5.4  Activated carbon—Filtrasorb—
 200. available from Calgon Corporation.
 Pittsburgh. PA.  or equivalent.
  SJ  Standards."
  5.3.1  Bromoform 9615—available
 from Aldrich Chemical Company.
  5.5.2  Bromodichloromethane 97%—
 available from Aldrich Chemical
 Company.
  5.5.3  Chlorodibromomethane—
 available from Columbia Chemical.
 Incorporated. Columbia. S.C.
  5.5.4  Chloroform 99%—available
 from Aldrich Chemical Company.
  5.9  Organic-free water—Organic-
 free water is defined as water free of
 interference when employed in the
 procedure described herein.
  5J.1  Organic-free water is generated
 by passing tap water through a carbon
 filter bed containing carbon. Change the
 activated carbon whenever the
 concentration of any trihalomethane
 exceeds ig/L
  SAZ  AMilUporeSuper-QWater
 System or its equivalent may be used -to
 generate organic-free deionized water.
  5.3.3  Organic-free water may also be
 prepared by boiling water for 15
 minutes. Subsequently, while
 maintaining the temperature at 90* C
 bubble a contaminant fiee inert ga«
 through the water at 100 ml/minute for
  • P«XMM hu own Ml«cted n the OKI jolwnt
for this aoaiyau aacauM it aiuUa. on all at tb«
column*, w«fl before aay at th« tzthaloratihaatj.
High «ttriu>ia» or laboratory ttmpcratuna in txcna
at nf our suk* ch» UM of Ihi» tolwit
unpractical For tlMM-rMMOL aftaflutiv* solvent!
an aeeanteble: howevci. the anaiyafauy
txpeneaee basefuw variances in the elunon ami
«f 
  c. Rapidly inject the aloholic standard
into the expanded area of the filled
volumetric Jlask. Remove the needle as
fast as possible after injection.
  d Mix aqueous standards by inverting
the flask three times only.
  & Discard the patents contained in
the neck of the flask. Fill the sample
syringe from the  standard solution
contained in the expanded area of the
flask as directed in Section 7.
  L Never use pipets to dilute ox transfer
samples and aqueous standards.
  g. Aqueous standards, when stored
with a headspace. are not stable and
should be discarded after one hoar.
Aqueous standards can be stored
according to Sections 6.4.9 and 7.2.
  5.9   Calibration standards.
  5.9.1  Prepare, from the standard
stock solutions, a multicomponent
secondary dilution mixture in methyl
alcohol so that a 20 pi injection into 100
ml of organic-free water will generate a
calibration standard which produces a
response close (± 25%} to that of the
unknown. (See 3.1.)
  5^-2  Alternative calibration
procedure.
  5.9-2.1   Construct a calibration curve
for each trihalomethane containing a
minimum of 3 different concentrations.
Two of the concentrations must bracket
each unknown.
  5.9.3  Extract and analyze  the
aqueous calibration standards in the
same  manner as the unknowns.
  5.9.4  Other calibration procedures
(7) which require the delivery of less
than 20 ill of methanolic standards to
10.0 ml volumes of water contained in
the sample syringe are acceptable only
if the  methanolic standard is delivered
by the solvent flush technique (S).
  5.10 Quality Check Standard
Mixture.
  5.10.1  Prepare, from the standard
stock  solutions, a secondary dilution
mixture in methyl alcohol that contains
10.0 cg/fil of each compound. (See 5.7.3
and 5.7.3.)
  5.1CL2  Daily, prepare and analyze a
2.0 fig/1 aqueous dilution from this
mixture by dosing 20.0 ul into 100 ml of
organic-free water (See Section 3.1).
  6. Sample Collection and Handling.
  3.1   The sample containers should
have a total volume of at least 25 mL
  3.1.1  Narrow-mouth screw-cap
bottles with the TFE fluorocarbon faced
silicone septa cap liners are strongly
recommended.
  8.2   Glassware Preparation.   	
  5.2.1  Wash all sample bottles, TFE
seals, and extraction flasks in detergent.
Rinse with tap water and finally with
distilled water.
  8^2  Allow the bottles and seals to
air dry, then place in an 105* C oven for
1 hour, then allow to cool in an area
known to be free of organic*.
  Not*.—Do aot heat the TEE seals for
extended periods of time (>1 hour) because
the siMcone layer slowly degrades at 105" C
  8.2.3  When cool, seal the bottles
using the TFE seals that will be used for
sealing the samples.
  9.3   Sample stabilization—A
chemical reducing agent (Section 5.3) is
added to all samples in order to arrest
the formation qf additional

-------
         Federal  Register / VoL 44. No. 231  / Thursday, November 29. 1979  /  Rules and Regulations   68685
trihalomelhanes after sample collection
(7,9) and to eliminate the possibility of
free chlorine reacting with impurities in
the extraction solvent to form interfering
organohalides. DO NOT ADO THE
REDUCING AGENT TO SAMPLES A T
COLLECTION TIME WHEN DATA
FOR MAXIMUM TRIHALOMSTHANE
FORMATION IS DESIRED. If chemical
stabilization is employed, then the
reagent is also added to the blanks. The
chemical agent (2.5 to 3 mg/40 mil is
added in crystalline form  to the empty
sample  bottle just prior to shipping to
the sampling site. If chemical
stabilization is not employed at
sampling time then the reducing agent is
added just before extraction.
   8.4  Sample Collection.
   6.4.1  Collect all samples in duplicate.
   6.4.2  Fill the sample bottles in such a
manner that no air bubbles pass through
the sample as the bottle is filled.
   6.4.3  Seal the bottle so that no air
bubbles are entrapped in it
   6.4.4  Maintain the hermetic seal on
the sample bottle until analysis.
   6.44  The raw source water sample
history should resemble the finished
drinking water. The average retention
time of the finished drinking water
within the water plant should be taken
into account when sampling the raw
source water.
   6.4.8   Sampling from a water tap.
   6.4.8.1  Turn on the water and allow
the system to flush until the temperature
of the water has stabilized. Adjust the
flow to about 500 ml/minute and collect
duplicate samples from the flowing
stream.
   6.4.7   Sampling from an open body of
water.
   6.4.7.1  Fill a 1-quart wide-mouth
bottle with sample from a representative
area. Carefully fill duplicate sample
bottles from the  1-quart bottle as in 6.4.
   8.4.3   If a chemical reducing agent
has been added  to the sample bottles.
fill with sample just to overflowing, seal
 the bottle, and shake vigorously for 1
minute.
   6.4.9  Sealing practice for septum seal
screw cap bottles.
   6.4.9.1  Open the bottle and fill to
overflowing. Place on a level surface.
Position the TFE side of the septum seal
upon the convex sample  meniscus and
seal the bottle by screwing the cap on
 tightly.
   6.4.9.2  Invert the sample and'lightly
 tap the cap on a solid surface. The
 absence of entrapped air indicates a
 successful seal. If bubbles are present
 open the bottle,  add a few additional
 drops of sample, then reseal bottle as
 above.
   6.4.10  Sample blanks.
  8.4.10.1  Prepare blanks in duplicate
at the laboratory by filling and sealing
sample bottles with organic-free water
just prior to shipping the sample bottles
to the sampling site.
  8.4.10.2  If the sample is to be
stabilized, add an identical amount of
reducing agent to the blanks.
  8.4.10.3  Ship the blanks to and from
the sampling site along with the sample
bottles.
  8.4.10.4  Store the blanks and the
samples, collected at a given site
(sample set], together in a protected
area known  to be free from
contamination. A sample set is  defined
as all the samples collected at a given
site (i.e.. at a water treatment plant
duplicate raw source water, duplicate
finished water and the duplicate sample
blanks comprise the sample set).
  8.5  When samples are collected and
stored under these conditions, no
measurable  loss of trihalomethanes has
been detected over extended periods of
time (7). It is recommended that the
samples be analyzed within 14  days of
collection.
  7. Extraction and Analysis.
  7.1  Remove the plungers from two
10-ml syringes and attach a closed
syringe valve to each.
  7.2  Open the sample bottle' (or
standard) and carefully pour the sample
into one of the syringe barrels until it
overflows. Replace the plunger and
compress the sample. Open the syringe
valve and vent any residue air  while
adjusting the sample volume to 10.0 mL
Close the valve.
  73  Fill the second syringe in an
identical manner from the same sample
bottle. This  syringe is reserved for a
replicate analysis (see 8.3 and 8.4).
  7.4  Pipette 2.0 ml of extraction
solvent into a dean extraction  flask.
  7.5  Carefully inject the contents of
the syringe into the extraction flask.
  7.8  Seal with a Teflon faced septum.
  7.7  Shake vigorously for 1 minute.
  7.3  Let stand until the phases
separate (/60 seconds).
   7.3.1  If the phases do not separate on
standing then centrifugation can be used
 to facilitate separation.
   73  Analyze the sample by  injecting
 3.0 pi (solvent flush technique. (8)] of the
 upper (organic) phase into the  gas
 chroma togra ph.
   8. Analytical Quality Control.
   8.1   A 2 p.g/1 quality check standard
 (See 5.10) should be extracted  and
 analyzed each day before any  samples
 are analyzed. Instrument status checks
   • If for any nason the chemical reducing agent
 has not been added to the temple, then it must be
 added just prior to analyses at the rate of 2J to 3
 mg/40 mi or by adding 1 mg directly to the sample
 in the extraction flask.
and lower limit of detection estimations
based upon response factor calculations
at 5 times the noise level are obtained
from these data. In addition, the data
obtained from the quality check
standard can be used to estimate the
concentration of the unknowns. From
this information the appropriate
standards can be determined.
  &2  Analyze the sample blank and
the raw source water to monitor for
potential interferences as described in
Sections 3.1.3.2, and 3.3.
  3.3  Spiked samples.
  8.3.1  For those laboratories
analyzing more than 10 samples a day,
each 10th sample analyzed should be a
laboratory-generated spike which
closely duplicates the average finished
drinking water in trihalomethane
composition and concentration. Prepare
the spiked sample in organic-free water
as described in section 5.9.
  8.3.2  In those laboratories analyzing
less than 10 samples daily, each time the
analysis is performed, analyze at least
one laboratory generated spike sample
which closely duplicates the average
finished drinking water in
trihalomethane composition and
concentration. Prepare the spiked
sample in organic-free water as
described in section 5.9.
  3.3.3  Maintain an up-to-date log.on
the accuracy and precision data
collected in Sections 3.3 and 8.4. If
results are significantly different than
those cited in Section 10.1, the analyst
should check out the entire analysis
scheme to determine why the
laboratory's precision and accuracy
limits are greater.
  8.4  Randomly select and analyze
10% of all samples in duplicate.
  8.5  Analyze all samples in duplicate
which appear to deviate more than 30%
from any established norm.
  8.6  Quarterly, spike an EMSL-
Gncinnati trihalomethane quality
control sample into organic-free water
and analyze.
   8.8.1  The results of the EMSL
trihalomethane quality control sample
should agree within 20% of the true
value for each trihalomethane. If they do
not the analyst must check each step in
 the standard generation procedure to
 solve the problem.
   &J   It is important that the analyst be
 aware  of the linear response
 characteristics of the electron capture
 system that is utilized. Calibration
 curves should be generated and
 rechecked quarterly for each
 trihalomethane over the concentration
 range encountered in the samples in
 order to confirm the linear response
 range of the system. Quantitative data
 cannot be calculated from non-linear

-------
68686    Federal Register  /  Vol. 44. No. 231  / Thursday.  November 29.  1979  / Rules and Regulations
responses. Whenever non-linear
responses are noted, the analyst must
dilute the sample for reanalysis.
  8.3  Maintain a record of the
retention times for each thhalomethane
using data gathered from spiked
samples and standards.
  8.3.1   Daily calculate the average
retention time for each thhalomethane
and the variance encountered for the
analyses.
  8.8.2   If individual thhalomethane
retention time vanes by more than 10%
over an eight hour period or does not fall
within 10% of an established norm, the
system is "out of control" The source of
retention data variation most be
corrected before acceptable data can be
generated.
  9.  Calculations.
  9.1  Locate each thhalomethane in
the sample chromatogram by comparing
the retention time of the suspect peak to
the data gathered in 3.8.1. The retention
time of the suspect peak must fall within
the limits established in 3.3.1 for a single
column identification.
  9.2  Calculate the concentration of
each trihalomethane by comparing the
peak heights or peak areas of the
samples to those of the standards.
Round off the data to the nearest jig/I or
two significant Figures.
Concentration, p.g/1 » sample peak height/
  standard peak height x standard
  concentration. >ig/L

  9.3  Calculate the total
thhalomethane concentration (TTHM)
by summing the 4 individual
thhalomethane concentrations in ^g/1:
TTHM (Mg/11 « (cone GHO»)-r(conc.
CHBrCUJ-Kconc. CHBr,Q]-Hconc.
CHBr,)
  9.4  Calculate the limit of detection
(LOO) for each trihalomethane not
detected using the following criteria:
                 lAXAJT(\
      LOO (*«/« -         1 < a H4/0
 Where:
 8 » peak height (nun) of 2 Mg/1 quality check
    standard
 A — 5 times the noise level in mm at the
    exact retention time of the
    trih* oaethane of the base line
    displacement in mm from theoretical
    zero at the exact retention time for the
    trihalome thane.
 ATT  •  attenuation factor.
  9-5  Report the results obtained from
the lower limit of detection estimates
along with the data for the samples.
  10.  Precision and Accuracy
  10.1  Single lab precision and
accuracy. The data in Table II were
generated by spiking organic-free water
with trihalomethanes as described in
5.9. The mixtures were analyzed by the
analyst as true unknowns.
                 Tims for Tnhtlomt!tian»t

                    Column  Column  Column
                      A      9      C
CMorofom
(OttroniaamraflMinmt
10
u
13

9.5
                              13     49
                             •tS    110
                              SS    31
                                                                       10.9
                                                                              384
                                           •On DM column. tncftforownyl*"* i common ra» saure*
                                           iMr egntiflMM. SOMIIM «*n Bromoacnioramcman*.
                     Taote IL—Stngtt Laboratory Accuracy ami Pneaon
                                                               »^oswn
                                 OOWWM*
                                           Nurno* o<
                                            urngxt
                                                     Mean m/i
                               Accuracy
                               a«rc«nt
Compound:
CHOi
CHfl^l*
rw*TJ*7!i
OH9f«CI
CHer>Cl
CMfr^
l^f*

91
88
.,„.., 1 i

2,7
	 ., 	 17



10
n
tj
19
2.0
16
13
16


13
96
1 4
17
9.9
10
12

110
106
10S
125
74
94

114

 Reference*
  1. Mieure. J. P.. "A Rapid and Sensitive
 Method for Determining Volatile
 Organohaiides in Water." Journal A WWA.
 69. 60.1977.
  2. Reding, R., et al. "TOM'S in Drinking
 Water Analysis by LLE and Comparison to
 Purge and Trap". Organics Analysis in Water
 and Wastewater. STP 838 ASTM. 1979.
  3. "Handbook for Analytical Quality
 Control in Water and Waste water
 Laboratories." Analytical Quality Control
 Laboratory, National £nvtronmentai
 Research Center. Cincinnati. Ohio. June 1972.
  4. Budde.  W. L. J. W. Schelberger.
 "Organic Analysis  Using Gas
 Chromatognphy-Mass Spectrametry." Ann
 Arbor Science. Ann Arbor. Michigan. 1979.
  5. "The Analysts of Trihalomethanes in
 Finished Water by  the Purge and Trap
Method,'* Environmental Monitoring and
Support Laboratory. Environmental Research
Center. Cincinnati  Ohio. 45288. May 15.1979.
  8. Richard J. f.: G. A. funk. "Liquid
Extraction for Rapid Determination of
Halomethanes in Water. Journal A WWA. S3
62. January 1977.
  7. Brass. H.}.. et al.. "National Organic
Monitoring Survey: Sampling and PurgeabU
Organic Compounds. Drinking Water Quality
Through Source Protection." R. B. Poiasek.
Editor. Ann Arbor Science, p.  398.1977.
  3. White. L. D.. et al. "Convenient
Optimized Method for the Analysis of
Selected Solvent Vapors in Industrial
Atmosphere," AIHA Journal. Vol. 31. p. 223.
1970.
  9. Kopfler. F. C. et aL "GC/MS
Determination of Volatiles for the National
Organics Reconnaissance Survey (NORS] or
Drinking Water. Idanuficanon and Analysis
of Organic Pollutants in Water." L R Keith.
Editor. Ann Arbor Science, p.  37.1978.

-------
       Federal Register / Vol. 44. No. 231 / Thursday. November 29. 1979 / Rules and Regulations   68687
         o
         VI
       u
       LU
                           COLUMN  PACKING:  3r.S?-10CO
                           CARRIES CAS: 57. ChU IN ARGON
                           CARRIER FLOW: 60 ML/MIN.
                           COLUMN TEMPERATURE:  50'C
                           DETECTOR:  ELECTRON CAPTURE
o
X
LU
Z
s
s
o
               a
               O
                     UJ
                     Z
                     uu
                     s
                     o
                     a
                     O
                     u
          RSTSNT1ON TIME IN MINUTES
FIGURE  1.  FINISHED   WATER  EXTRACT

-------
58888
Federal Register / VoL 44. No. 231 / Thursday. November 29. 1979 / Rules and Regulations













^>

o
Uk
O L
« £
n -
-j c
COLUMN PACKING: 10%
SQUALANE CARRIER
FLOW: 25m!/min COLUMN
!£ TEMPERATURE: 67
Z LU
< Z
? 5
^^? ^^^
< u^
0 S
« o
O 2
= o
u «
TT C3
a —
i o 2
_ fi£ Of
_l ri **
^22 2
— — , •" Lk>
- e s n
= O
G —
512X


I
i






(j
<
fi£
>-
Li
^

w

Z
o
02
*i
!l024X

1





M 1
<
u
ft
Ll
t ' '

0 1 2
11


w /
i
o
Q£
C3
256X1
i


1
1


1

I K
i t i t i t r F '
* i t i • i i i • r
3456 7 3 9 10 11 12
                       TIME (min)



       FIGURE  2. EXTRACT  OF  STANDARD

-------
     Federal Register / Vol. 44. No. 231 / Thursday. November 29. 1979 / Rules and Regulations   68689

                 COLUMN PACKING:   6% OV-11 + 4% SP-2100
       vu              CARRIER FLOW:   25 ml/min
       -    TEMPERATURE  PROGRAM:   45°C-12 MINUTES
     .  o                              1°/M1NUTE   TO  70°C
       ^
       u



^^M
LOROFORA/
^•w
/

1

Of
UJ
h-
r\ "^
5 S < z "
< ^-- — UJ 2
UOROETH
1
»
«


!

^% ^M 1 ^
^J uj >•
•^ O ^" ua
' i 1 II
o 5 C5g
cs o < —
0 2 « g
52 SS • 2
^^ ^^ i O£
£ i = 2
IT 1
i
i a
\ll 1 A

    5     10    15     20   25    30   35    40    45
                     TIME (min)

      FIGURE 3, EXTRACT OF  STANDARD
SIUJNO cooe ueo-01-e

-------
68690   Federal  Register /  Vol. 44.  No. 231 / Thursday. November  29. 1979 / Rules  and Regulations
Part III—Determination of Maximum
Total Trihalomethans Potential (MTP)
  The water sample used for this
determination is taken from a point in
the distribution system that rejects
maximum residence time. Procedures for
sample collection and handling are
given in EMSL Methods 301.1 and 39-1.2.
No reducing agent is added to "quench"
the chemical reaction producing THMs
at the time of sample collection. The
intent is to permit the level ofTHM
precursors to  be depleted and the
concentration of the THMs to be
maximized for the supply being tested.
  Four experimental parameters'
affecting maximum THM production are
pH. temperature, reaction time and the
presence of a  disinfectant residual.
These parameters are dealt with as
follows:
  Msasure the disinfectant residual at
the selected sampling point Proceed
only if a measurable disinfectant
residual is present. Collect triplicate 40
ml water samples at the pH prevailing at
the time of sampling, and prepare a
method blank according to the EMSL
methods. Seal and store these samples
together for 7  days at 25"C or above.
After this time period, open one of the
sample containers and  check for
disinfectant residual. Absence of a
disinfectant residual invalidates the
sample for further analyses. Once a
disinfectant residual has been
demonstrated, open another of the
sealed samples and determine total
THM crncentratioa using either of the
EMSL analytical methods.
                                                                   U.S.  Dr,i
                                                                   Chicago, Uiinoia
                                                                                      S.  *m C—

-------