EPA-600/4-75-002
        LOW COST COMPACT
 X-RAY FLUORESCENCE  ANALYZER
    FOR ON-SITE  MEASUREMENT
OF TRACE  ELEMENTS IN AIRBORNE
      PARTICULATE EMISSIONS
                   by

           L . S . Birks and J . V . Gilfrich

            Naval Research Laboratory
             Washington, D. C. 20375
       Interagency Agreement No. EPA-IAG-D4-0344
              ROAP No. 26AAN-012
            Program Element No. 1AA010
         EPA Project Officer: Dr. Jack Wagman

       Environmental Sciences Research Laboratory
          Office of Air, Land, and Water Use
       Research Triangle Park, North Carolina 27711
                Prepared for

      U.S. ENVIRONMENTAL PROTECTION AGENCY
         Office of Research and Development
             Washington, D. C. 20460

                 July 1975

-------
                         EPA REVIEW NOTICE

This report has been reviewed by the National Environmental Research
Center - Research Triangle Park, Office of Research and Development,
EPA,  and approved for publication.  Approval does not signify th^t the
contents necessarily reflect the views and policies  of the Environmental
Protection Agency, nor does mention of trade names or commercial
products constitute endorsement or recommendation for use.
                    RESEARCH REPORTING SERIES

Research reports of the Office of Research and Development, U.S. Environ-
mental Protection Agency, have been grouped into series.  These broad
categories were established to facilitate further development and applica-
tion of environmental technology.  Elimination of traditional grouping was
consciously planned to foster technology transfer and maximum interface
in related fields.  These series are:

           1 .  ENVIRONMENTAL HEALTH EFFECTS RESEARCH

           2 .  ENVIRONMENTAL PROTECTION TECHNOLOGY
           3.  ECOLOGICAL RESEARCH
           4.  ENVIRONMENTAL MONITORING

           5.  SOCIOECONOMIC ENVIRONMENTAL STUDIES

           6.  SCIENTIFIC AND TECHNICAL ASSESSMENT REPORTS
           9.  MISCELLANEOUS

This report has been assigned to the ENVIRONMENTAL MONITORING
series.  This series describes research conducted to develop new or
improved methods and instrumentation for the identification and quanti-
fication of environmental pollutants at the lowest conceivably  significant
concentrations. It also includes stuclii-s to determine the ambient concen-
trations  of pollutants in the environment and/or the  variance of pollutants
as a function of time or meteorological factors.
                                 \
This document is available to the public for sale through the National
Technical Information Service, Springfield, Virginia 22161.

                 Publication No.  EPA-600/4-75-002
                                  11

-------
                           CONTENTS

Abstract „ ..... » .......... . „ o . . .... . .... . . „ . „ . . <> <, .... . . . . . . ,, .   iv
Problem Status . . .	.........o...	.................o   iv
Authorization  ..............................................   iv
DNTTRODUCT ION  ...........................................   1
COMMERCIAL INSTRUMENTS	   1
INSTRUMENT DESIGN ......................................   2
RESULTS
   Sensitivity and Detection Limits	   7
   Sample Measurement ....................................   9
   Field Test	   9
SUMMARY .................................................   10
ACKNOWLEDGMENT .......................................   11
REFERENCES  ...............	.................	.	   11

FIGURES
   1.  Block Diagram of Compact X-Ray Analyzer .............   3
   2.  Plan view of spectrometer  ............................   5
   3.  Photograph of the front panel of the
   Compact X-Ray Analyzer as delivered	   6
   4.  Photograph of spectrometer with sample
   chamber open ...........................................   6
   5.  Analyst inserting sample  .............................   7
   6.  Detection Limits as a function of atomic number	   8
                              111

-------
                            ABSTRACT

    Alow cost, compact wavelength dispersion x-ray fluorescence
analyzer has been designed using mostly state-of-the-art commercial
components.  A prototype instrument has been built for less than
$5000 and tested  in the field.  With x-ray tube power of 80 watts
(40 kV, 2 mA) the 100 second 3
-------
                           INTRODUCTION

    Fossil-fueled power plants,  incinerators, and other sources emit
substantial quantities  of airborne particulate matter containing various
amounts of such elements as Pb, Hg, Cd,  As, Ni,  V, Cr,  Se,  Mn, Cu
and Sb that are  of concern as actual or potential hazardous pollutants.
An effective and economical means for on-site monitoring  of emissions
of these elements would be of considerable help in developing control
strategy.   The technique should be simple, rapid and have limits of de-
tection adequate to analyze fairly short-time collections with an accept-
able accuracy.  The necessary equipment  should be inexpensive so that
even small operators  could afford the capital expenditure.

    X-ray  fluorescence had already been demonstrated to be a valuable
analytical  tool for the measurement of the  elemental composition of
particulate pollution.'    '  In view of this fact,  the U.  S.  Environ-
mental Protection Agency (EPA) requested the Naval Research .Labor-
atory (NRL) to suggest a commercial x-ray analysis system to perform
the monitor function,  or to design and build an instrument  specifically
suited to this  application.
                   COMMERCIAL INSTRUMENTS

    The problem of evaluating low-cost x-ray instrumentation is com-
plicated by the degree of sophistication which has been introduced by
the x-ray manufacturers.  The cost per sample has been reduced using
this expensive equipment for those situations where the sample  load is
high enough to take fall advantage  of the automated instruments. How-
ever, the capital expenditure is impossible to justify for any industry
which needs to analyze only a few  samples»

    Fortunately,  there are  several inexpensive x-ray analyzers com-
mercially available and  our first step was to examine their  capabilities
using a set of criteria agreed upon between the authors and  the EPA
project officer.  It was felt that a  reasonable target cost for the instru-
ment was $5000; it should  have a  100 second detection limit of about
          for a single element (as particulate material deposited on a


                                  1

-------
thin substrate);  it should be capable of measuring any one of a wide
variety of elements witha resolution adequate to minimize interferences;
it should be  simple to use,  compact and able to operate from a conven-
tional 110 volt, 60 Hertz A.C.  branch  circuit.   The types of instruments
considered and the  factors  involved in  evaluating them are listed in
Table I.
TABLE I.   PROPERTIES OF COMMERCIAL LOW COST X-RAY ANALYZERS

                                            Solid State
                                            Det. -X-Ray Tube
Type of
Instrument
Cost
Weight
Resolution
Scint. or Prop.
Det. (jv1 Filters
-$5000
20 Ibs
Varies (\?
Solid State
Det. -Isotope
~ $8000
100 Ibs
180 eV @ 4 k<
 100 Second

 Det. Limits


 Special

 Requirements
            each filter

            pair used.

            Typically 500 eV


            Few to Few

            Tens of fig/cm


            Need to examine

            each element to

            determine if approp.

            filters available.
ZOO eV @ 6 keV
220 eV @ 10 keV
270 eV @ 20 keV

< 1 Hg/cm
Liq. N-, (can

have 28 day

holding time)
~ $15000

285 Ibs

180 eV @ 4 keV

200 eV @ 6 keV

220 eV @ 10 keV

270 eV @ 20 keV


< 0. 5 fig/cm



Liq. N-, (can

have 28 day

holding time)
                                                            Crystal

                                                            Spectrometer


                                                            ~ $10000

                                                            125 Ibs

                                                            25eV@4keV

                                                            68 eV @ 6 keV
                                                            143 eV @  10 keV

                                                            650 eV @ 20 keV
< 1 fig/cm
                               None
                                                                    2
    As can be seen from Table I, none  of the commercial instruments
satisfy all the criteria.  The first instrument is the only one which
satisfies the cost criterion and is the only one which does not meet the
detection limit  requirements. The  most discouraging observation from
the table is the poor energy resolution of all of these particular systems.
The lack of good energy resolution results  in significant interferences
which only can  be  overcome for the solid-state detector instruments by
computer processing of the  da,ta  (one of the ways in which the modern
equipment has become  sophisticated arid expensive);  the instruments
using filters are timply not suited  to any analysis situation involving
such interferences.  The particular crystal spectrometer listed used a
low-power x-ray tube and inefficient geometry which made it necessary
to sacrifice  resolution  for intensity,, In general, however,  the  resolu-
tion of a crystal spectrometer can  be considerably better than that
listed in the table  and suggested that such an instrument should be de-
signed and constructed  specifically  for the  problem at hand.

                       INSTRUMENT DESIGN

    It was  not the intent of the eifort reported here to perform research
in the area of crystal spectrometer design.  Rather the instrument
would use  those state-of-the-art components which were commercially
available and we would  only  custom-make those parts which we re unique.

-------
The Interagency Agreement between NRL, and EPA specified the con-
struction of a bread-board model but it was recognized early that a
prototype would be much more useful.  The complexity introduced by a
vacuum system seemed unnecessary for this prototype and therefore
the instrument would be limited to measuring those elements whose
characteristic x-rays are not attenuated significantly in an air path.
               BEAM
              ,TRAP
SHIELDED
 SAMPLE
CHAMBER
      SAMPLE
      X-RAY
      TUBE
                                           PULSE HEIGHT
                                             ANALYZER
                     FUSED MAIN POWER SWITCH
Figure  1.   Block diagram of the Compact X-Ray Analyzer.

    Figure 1 is a block diagram of the complete x-ray analysis  system
and Table II lists the commercial components and their cost, as well
as the parts  fabricated at NRL with the best estimate (probably good to
10 or 20%) of the cost to produce the latter.  Assembly and testing of
the equipment could be easily  accomplished for an additional thousand

-------
dollars,  although a considerably larger amount was spent on the proto-
type to insure that the approach used was the most appropriate.
                             TABLE  II

          COMPONENTS FOR COMPACT X-RAY ANALYZER

                   PURCHASED COMMERCIALLY
Regulator
X-Ray Tube H.V. Supply
X-Ray Tube
"005 X 4" Collimator
Scintillation Detector
Detector H. V. Supply
Recorder
Electronics Power Supply
                           Manufacturer and Model No.
                           Sola 23-22-150
                           Universal-Voltronics BPE-40-4
                           Machlett EG-50 HW
                           Philips 19011200
                           Harshaw SHG
                           Venus Scientific K-15
                           Rustrak 288
                           Sorensen MMD-12. 120
                                                               Cost
$
Crystal (LiF)

Filament Supply
X-Ray Tube Housing
  and Sample Chamber
Spectrometer,  incl.
  Crystal and Detector
  Mounts
Counting Electronics:
  Amp., PHA, Ratemeter
  Sealer,  Timer, etc.
                       FABRICATED AT NRL

                           Cleaved, abraded and etched,
                            from NRL stock
                           6.3 VCT Transformer, 1.5Amp
                            Variac and miscellaneous
                            small parts
                           Manufactured  in NRL Shops:
                            materials and labor
                           Gears ($36) +  machine work
                            in NRL Shops:  Mat'ls and labor
                           Some circuitry designed in
                            NRL Electronic Shop, other
                            circuitry standard: Mat'ls
                            and labor
Miscellaneous Cables
 Connectors and
 Electronics Comp.
Cabinet to house components fabricated at NRL
 100
 415
1050
 255
 310
 150
 197
  96
$ -50
  -25
  •500

  •100


  -500



  ~25


  -50
                                                     TOTAL   $3823

    The Sola regulator was used to stabilize the line voltage.  The
Universal-Voltronics  high voltage supply was  chosen largely on the
basis of cost and size. Although a 50 kV supply would have been pre-
ferable in order to excite the heavier elements, Universal Voltronics
had not built this power supply for more than  30 kV; therefore they
were only willing to upgrade it to 40 kV, which we considered acceptable.
The Machlett EG series x-ray tube was the  only low-powered tube
readily available; it was purchased with a tungsten target for  efficient
excitation over a wide atomic number range and with an unnecessarily

-------
thin window ("„ 005 Be) because that particular model was in stock for
rapid delivery.  This x-ray tube is located in an oil-filled housing
which, together with the specimen chamber, was fabricated at NRL..
The tube housing is air cooled and contains a resistor stack by •which
the voltage applied to the tube  is monitored.  The specimen chamber
contains a sample changing mechanism, the beam trap to prevent scat-
tering of the primary beam which is transmitted through the sample,
and the collimator.  The x-ray tube current is controlled by the  NRL-
built filament supply.  Both the x-ray tube housing and the  specimen
chamber were lined with lead to provide radiation protection.

    The spectrometer, consisting of the crystal and detector and the
6-29 coupling  between them by means of a planetary gear  arrangement,
is shown in Figure 2.  The Harshaw SHG scintillation detector has a
very good reputation with regard to its energy resolution and low noise.
The detector high voltage supply,  the  low voltage power supply for the
electronics package, and the recorder were commercial units; the  rest
of the  counting electronics were designed and built at NRL,  making use
of modern conventional circuitry.  Figures 3 through 5 are photographs
of the  completed instrument.
                SAMPLE
               CHAMBER

                            COLLIMATOR

                                       CRYSTAL
                                              SCINTILLATION
              X-RAY  TUBE              \  \  DETECTOR
            (BELOW SAMPLE)
Figure 2,  Plan view of spectrometer.

    The operation of this prototype is identical with that of any wave -
length dispersive x-ray fluorescence analyzer with which it  is assumed
the reader is familiar.  The spectrometer is manual and must be set
by hand for detecting the x-rays of interest;  the  crystal is coupled to
the detector and automatically set at the diffracting angle.  The in-
tensity can be read on the sealer, ratemeter  or strip-chart  recorder.

-------
Figure 3.  Photograph of the front panel of the Single Element
Analyzer as delivered.
Figure 4,  Photograph of the spectrometer with the sample chamber
open showing a sample in place.

-------
Figure 5.  Analyst inserting sample into instrument (preliminary
front panel).
                             RESULTS

    Sensitivity and Detection Limits

    One of the  parameters considered important in defining the useful-
ness of the compact x-ray analyzer described here is the detection
limit for various elements.  The original criterion was that the 100
second detection limit should be  1 /Jg/cm^ or better for the  elements
within its wavelength  range.  Measurements  of the sensitivity
[ (c/100 s)/(/J,g/cm2)]  and the detection limits (jJg/cmZ) were made
using single element standards of solutions deposited  on filter paper.(^)
The results of these measurements are  listed in Table III and the
detection limits are shown as a function of atomic number in Figure 6.
It can be seen  that the 1 ^tg/cm^  criterion is satisfied for all elements
tested with the exception of Cd which is  1.2 jig/cm20  Many of the
elements  demonstrate detection limits significantly better, the best
being Zn at 0.  1 fig

-------
                          TABLE III
             SENSITIVITY AND DETECTION LIMITS*
Element X-Ray

V Ka
Fe Ka
Cu Ka
Zn Ka
Br Ka
Mo Ka
Cd Ka
W La
Pb La
U La
Line Sensitivity
[(c/100s)/(ng/cm2)]
245
1249
1582
2111
1160
596
98
587
714
553
BCD
(c/lOOs)
1429
3659
7974
5424
7009
6255
1504
34586t
7993
6480
100 second measurements with W target Machlett EG-50
tube operated at
High background
10 -
100
40 kV, 2 mA.
due to scattered primary W

SECOND DETECTION LIMITS

L lines.


3a D.L.
(Mg/cm2)
0.46
0. 15
0. 17
0. 10
0.22
0.40
1.2
0. 95
0.38
0.44
x-ray




W TARGET EG-50 (40kV, 2mA)

^^,
"E
o>
Ji.
1-
I i.o-
-z.
o
" \
t \
Q \

n 1 1 	 £D_
O Ka LINES
X La LINES



/
/
O

/
O/
/
1 1 1 1





^
\
\ 	 	
X










-^x




        '20
40           60
 ATOMIC  NUMBER
80
Figure 6.  Detection limits as a function of atomic number.

-------
    Sample Measurement

    The instrument was further tested by analyzing some auto exhaust
samples provided by EPA.  During these measurements,  some diffi-
culty was experienced due to a faulty component in the detector high-
voltage supply.  Unfortunately, these samples had also been analyzed
in the standard laboratory spectrometer and had received enough
radiation exposure that they were very brittle.  While attempting to
remeasure these samples in the single element analyzer they deterio-
rated and only a few  of the  samples could be analyzed for some of the
elements. The  results of these  few determinations are shown in Table
IV.  The  1'less than" value  of 0.  13 |J,g/cm2 listed for  Br  in sample 07
is not inconsistent with the detection limits shown in Table III. Sample
07 was collected on a Nuclepore filter which, because it is significantly
thinner than  Millipore or Whatman filters,  contributes a smaller scat-
tered background.  This leads to better detection limits than those
listed in Table in (0. 13 fig/cm2 for Br compared to 0.22 jig/cm2 in
the Table).

                              TABLE  IV

          COMPARISON OF COMPACT X-RAY ANALYZER

              RESULTS WITH THOSE OBTAINED USING

          A  STANDARD LABORATORY X-RAY ANALYZER

                               Concentration, fig/cm

       Sample No.        05       06        07        08

       Pb
          PW 1410*       3.6      0.45
          CXA            5.2      0.50

       Br

          PW 1410       1. 1                0.07
          CXA            1.1               <0.13t     0.28

       Fe
          PW 1410               32                   4.7
          CXA                    34                   5.2
       >\<
        PW  1410  is the standard laboratory instrument.
        CXA is the compact x-ray analyzer.
       •f
        "Less than" value indicates the 3(T,  100  s. detection limit.

    Field  Test

    Since  the instrument was designed to be  operated  conveniently on
the site of an emission source, the final test before transferring it to
the EPA was conducted in the  field.  Arrangements were made by the
EPA to collect samples  of particulate material from one of the stacks

-------
at the River Bend Plant of the Duke Power Company in Charlotte,
North Carolina, and to measure several elements at that location
using the  compact x-ray analyzer.  The instrument was transported
from NRL to the EPA Environmental Sciences Research Laboratory in
Research Triangle Park, North Carolina by EPA vehicle. At that
location,  the analyzer was set up and put into operation to insure  that
it had not been damaged during  transport.  Some minor malfunctions
were remedied and the analyzer was made ready for the trip to
Charlotte, again in an EPA truck.  The original intention was to make
the measurements in the EPA instrument  shack located on the  roof of
the plant, immediately adjacent to the  stack fitted with sampling ports.
However, it was apparent that conveying the  instrument to the  roof
would not demonstrate  anything that could not be shown as well by
setting it  up in the plant laboratory and there was  some chance of
damage in attempting to navigate the tortuous path to the  roof.  As at
the EPA laboratory, the analyzer was  set up and put into operation.
Sample deposits on Millipore filters were taken from the stack and
measured for six elements using  "solution on filter paper" standards
prepared  at NRL.  The results  are shown in  Table V.   At the conclu-
sion of this operation,  the analyzer was transported back to the Environ-
mental Sciences Research Laboratory.

                               TABLE V

               COMPACT X-RAY ANALYZER RESULTS

            ON POWER PLANT PARTICULATE SAMPLES

                                Concentration
                              Sample  X-2   Sample X-5

               Mn                  0.4         0.5
               Fe                 31           33
               Ni                  0.3         0.5
               Zn                <0.8*        7.1
               Hg                  5.7         2.7
               Pb                  4.7          5.1

                Zn was not detectable on sample X-2.   The
               anomalously high 100 sec.  detection limit
               (0. 8 Jig/cm^) was  caused by the high Zn
               content of the "o-ring" used to clamp the
               filter in the sample holder.


                             SUMMARY   '

    The design, construction and  testing of the  instrument described in
this report demonstrates that a compact,  low powered, relatively in-
expensive, wavelength dispersion x-ray analyzer can be produced
using state-of-the-art components. It is capable of measuring ele-
mental composition of the particulate material collected from emission


                                  10

-------
sources (or from ambient air,  for that matter) with 100 second 3cr  de-
tection limits ranging from 0. 1 to 1 j^g/cm^ for elements which can be
analyzed in an air path instrument.  The cost of the components used
to fabricate such an instrument was well below $5000  and we estimate
that,  even in the present (1975) economy, if produced in  sufficient
quantity,  similar instruments should be commercially marketable for
not more than $ 10, 000.

    If an instrument of this type were to be  commercially produced as
a "single element" analyzer, this would enable the device to be cus-
tomized, e.g.,  the use of a (220) LiF crystal and 50 kV power supply
for measurements of the K lines of elements from Ru to  Cd; the am-
plifier gain,  PHA settings and Bragg angle  fixed  rather than adjustable;
a selection made of digital or analog readout depending on the applica-
tion,  and the sample  chamber specifically adapted to the  samples being
measured.  Thus the instrument might have only two controls, "on-off"
and "read", for simplicity of operation.   Conversely, the degree of
complexity could be increased to make the analyzer more versatile
(but unfortunately more  expensive)  by providing a motor drive for the
spectrometer,  enclosing the  specimen chamber and spectrometer in a
vacuum housing, installing a crystal changer and  a thin-window pro-
portional detector, etc.

    Several of the x-ray equipment  manufacturers have seen this  device
and discussed the design parameters involved in the production of an
inexpensive instrument of this type.  They all expressed  interest and
some of them are currently considering the possibility of making such
an instrument commercially.
                       ACKNOWLEDGMENTS

    The authors wish to express grateful appreciation to M. C. Peckerar,
N.  L.  Midkiff and J. H. McCrary of the X-Ray Optics Branch, Naval
Research Laboratory,  for assistance in various phases of the design,
construction and testing of this instrument.


                           REFERENCES

1.  "Development of X-Ray Fluorescence Spectroscopy for Elemental
Analysis of Particulate Matter  in the Atmosphere and in Source
Emissions, " L. S. Birks, J. V.  Gilfrich and P. G. Burkhalter,
EPA-R2-72-063, Oct.  1972.

2.  "Development of X-Ray Fluorescence Spectroscopy for Elemental
Analysis of Particulate Matter  in the Atmosphere and in Source
Emissions.  Phase II:  Evaluation of Commercial Multiple  Crystal
Spectrometer Instruments, " L,. S. Birks and J. V. Gilfrich, NRL,
Report 7617, June  1973.
                                 11

-------
3.  J. V.  Gilfrich,  P. G. Burkhalter and L. S. Birks, Anal. Chem.
45, 2002 (1973).

4.  F.  S.  Goulding  and J. M.  Jaklevic, Lawrence Berkeley Laboratory,
UCRL-20625,  May  1971.

5.  H.  R.  Bowman, J. G. Conway and F. Asaro,  Environ. Sci.
Technol. 6_, 558 (1972).

6.  T.  R.  Dittrich and C. R.  Cothern, J. Air Pollut.  Contr. Ass. 21,
716 (1971).

7.  To  B.  Johansson, R. Akselsson and S. A. E.  Johansson, Lund
Institute of Technology, LUMP 7109, Aug.  1971.

8.  R.  L.  Watson,  J. R. Sjurseth and R.  W. Howard, Nucl. Instrum.
Methods ^3, 69 (1971).

9.  J. S. Cooper, Nucl.  Instrum. Methods  106, 525 (1973).

10.  J.  Leroux and  M. Mahmud,  J.  Air Pollut. Contr. Ass. 20, 402
(1970).

11.  P.  Greenfelt,  A. Akerstrom and C.  Grosset, Atmos. Environ. J5,
1 (1971).

12.  C.  L. Luke, T.  Y.  Kometani, J. E. Kessler, T. C. Loomis,
J. L. Bove and B.  Nathanson, Environ Sci. Technol.  6,  1105 (1972).
                                 12

-------
                                    TECHNICAL REPORT DATA
                             (Please read Instructions on the reverse before completing)
 1. REPORT NO.
   600/4-75-002
                              2.
                                                             3. RECIPIENT'S ACCESSIOC+NO.
 4. TITLE AND SUBTITLE
   Low Cost Compact X-ray Fluorescence Analyzer  for
   On-Site Measurement of Trace Elements in Airborne
   Particulate Emissions	
                                                             5. REPORT DATE
                                                               July 1975
             6. PERFORMING ORGANIZATION CODE
 7. AUTHOR(S)
   L.  S. Birks and  J.  V. Gilfrich
                                                             8. PERFORMING ORGANIZATION REPORT NO.
9, PERFORMING ORGANIZATION NAME AND ADDRESS

   Naval Research  Laboratory
   Washington, D.  C.   20375
              10. PROGRAM ELEMENT NO.
               1AA010  (26AAN-012)
              11. CONTRACT/GRANT NO.

               EPA-IAG-D4-0344
 12, SPONSORING AGENCY NAME AND ADDRESS
   Environmental Sciences Research  Laboratory
   Office of Research and Development,  EPA
   Research Triangle Park, N.C.   27711
                                                             13. TYPE OF REPORT AND PERIOD COVERED
                                                              Final  - 2 yrs ending  6/75
              14. SPONSORING AGENCY CODE
 15. SUPPLEMENTARY NOTES
 16. ABSTRACT

        A low cost,  compact wavelength dispersion x-ray fluorescence analyzer
  has been designed using mostly state-of-the-art  commercial components.  A
  prototype instrument has been built for less than  $5000 and tested in the
  field.  With x-ray tube power of  80 watts (40 kV,  2  mA) the 100  second 3
-------