EPA-810/B-92-005
                      United States           Office of Water (WH-550)     EPA 810-B-92-005
                      Environmental Protection   Office of Pesticides and      February 1992
                      Agency               Toadc Substances (H-7501C)
                     QUALITY ASSURANCE PROJECT PLAN
                                     FOR THE
          NATIONAL PESTICIDE SURVEY OF DRINKING WATER WELLS
                ANALYTICAL METHOD 5 - METHYL CARBAMATES
                                    Prepared by:

                           Michael G. Winslow and David H. Greer
                         Environmental Science & Engineering, Inc.
                                    P.O. Box 1703
                                 Gainesville, PL 32602


                                    Prepared for:

                           U.S. Environmental Protection Agency
                               Technical Support Division
                                Office of Drinking Water
                              26 W. Martin Luther King Drive
                                Cincinnati, Ohio 45268
                                               US Environmental Protection Agency

                                                     5, Library (PL-12J)
                                                     iJackson Boulevard, 12th Floor
                                                          60604-3590

-------
                                                                Section No. 1
                                                                Revision No. 2
                                                                Date: July 29, 1990
                                                                Page 2 of 2
Paul Geiszler
R. Kent Sorrell
Portia Pisigan
Lora Johnson
                          APPROVAL PAGE
                           , ESE Project Leader
                           , TSD Technical Monitor
                           , ESE QAC
                           , NPS QAO
Elizabeth Leovey
                           , OPP QA Officer

-------
                                                                                                               Section No. 2
                                                                                                               Revision No. 2
                                                                                                               Date:  July 29, 1990
                                                                                                               Page 1 of  2
•>t-
<0
-•4
NATIONAL PESTICIDE SURVEY
QUALITY ASSURANCE PROJECT PLAN FOR
ANALYTICAL METHOD 5 - METHYL CARBAMATES
2. TABLE OF CONTENTS
Section
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
TITLE AND APPROVAL PAGE
TABLE OF CONTENTS
PROJECT DESCRIPTION
PROJECT ORGANIZATION AND RESPONSIBILITIES
QUALITY ASSURANCE OBJECTIVES FOR
MEASUREMENT DATA
SAMPLING PROCEDURES
SAMPLE CUSTODY
CALIBRATION PROCEDURES AND FREQUENCY
ANALYTICAL PROCEDURES
DATA REDUCTION, VALIDATION REPORTING
INTERNAL QUALITY CONTROL CHECKS
PERFORMANCE AND SYSTEM AUDITS
PREVENTIVE MAINTENANCE
SPECIFIC PROCEDURES FOR ASSESSING
MEASUREMENT SYSTEM DATA
CORRECTIVE ACTION
QUALITY ASSURANCE REPORTS TO
MANAGEMENT
ARCHIVAL OF RAW DA - ,
Pages
2
2
1
2
2
4
1
1
1
1
3
1
1
1
2
1
2
Revisions
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
Date
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90
7/29/90

-------
                                                                        Section No. 2
                                                                        Revision No. 2
                                                                        Date: July 29, 1990
                                                                        Page 2 of 2
2. TABLE OF CONTENTS (continued)
Appendices
A.





B.

C.
D.

E.
F.
G.
H.
1.
J.

METHOD 5. MEASUREMENT OF
N-METHYLCARBAMOYLOXIMES AND
N-METHYLCARBAMATES IN
GROUND WATER BY DIRECT
AQUEOUS INJECTION HPLC WITH
POST COLUMN DERIVATIZATION
NPS ANALYTE REPORTING BELOW MRL AND
IDENTIFYING UNKNOWN PEAKS
CHANGES IN NPS LABORATORY PROCEDURES
NPSIS SAMPLE RECEIPT SOFTWARE FOR
LABORATORIES
FORMAT FOR NPS DATA
DATA REPORTING FORMAT CHANGES
DATA REPORTING CODES
NPS RAPID REPORTING SYSTEM
DIXON'S TEST
STANDARD OPERATING PROCEDURE: BATCH
Pages Revisions Date





27 2 7/29/90

5 2 7/29/90
3 2 7/29/90

17 2 7/29/90
6 2 7/29/90
3 2 7/29/90
4 2 7/29/90
1 1 2 7/29/90
5 2 7/29/90

FILING SYSTEM
7/29/90

-------
                                                                             Section No  3
                                                                             Revision No. 2
                                                                             Date: July 29, 1990
                                                                             Page 1 of 1
3.    PROJECT DESCRIPTION
      The National Pesticide Survey is intended to assess the extent and nature of pesticide presence
in well waters used in private and community water supply systems.  The statistical design of the
survey will suggest from a manageable number of samples and analyses the nature of pesticide
presence in such water supplies throughout the nation.
      Environmental Science and Engineering, Inc. (ESE) of Gainesville, Florida, has contracted with
the U.S. Environmental  Protection Agency (EPA) to analyze the collected water samples for
carbamates (Method 5) and halogenated fumigants (Method 7).  This project plan applies to Method
5, the determination of aldicarb, aldicarb sulfone, aldicarb sulfoxide, baygon, carbaryl, carbofuran, 3-
hydroxy-carbofuran, methiocarb, methomyl, and oxamyl in ground waters.  Method 5 involves direct
injection of an aliquot of the water sample, high-pressure liquid chromatographic separation,
derivatization, and fluorescence detection and quantitation. Positive results will be confirmed by
second column analysis.  Method 5 is an adaptation of EMSL Method 531.1, "Measurement of N-
Methylcarbamoyloximes and N-Methylcarbamates in Water by Direct Aqueous Injection HPLC with
Post Column Derivatization"  (Appendix A).

-------
                                                                           Section No. 4
                                                                           Revision No. 2
                                                                           Date: July 29, 1990
                                                                           Page 1 of 2
4.    PROJECT ORGANIZATION AND RESPONSIBILITIES
     The project staff is specified in ESE's proposal to perform the contract. The project organization
chart is given in Exhibit 4-1.
     The managing staff are:
           Paul C. Geiszler             Project Director
           Michael G. Winslow          Project Manager and Technical Contact
           Portia O. Pisigan             Quality Assurance Coordinator
           Virgina C. O'Brien            Data Management and Sample Control
           Bradley A. Weichert          Manager, GC/HPLC Department
           W. Scott Keeran             Associate Scientist,  Lead Chemist
Certain specific project assignments are handled by others:
           Vince Prem-Das             Sample Check-in
           D.  Michael Ritter             Computer Sample Check-in and Data File Generation
           Scott Poole                  Sample Custody and Coldroom Support
     Samples will generally be received by Vince Prem-Das, ESE,  14220 Newberry Road, Gainesville,
FL, 32607; phone (904) 332-3318. It will generally be necessary to ask the ESE switchboard to page
Mr. Prem-Das.  In his absence, ask for Michael G. Winslow, Project Manager.
     The EPA Technical Monitor (primary contact) for Method 5 matters is
R. Kent Sorrell, phone (513) 569-7943. The EPA Project Officer is David  J. Munch.

-------
                                                  Section No. 4
                                                  Revision No. 2
                                                  Date: July 29, 1990
                                                  Page 2 of 2
                   EXHIBIT 4-1

          PROJECT ORGANIZATION CHART
            EPA TECHNICAL MONITOR
                    K. Sorrell
             ESE PROJECT DIRECTOR
                    P. Geiszler
             ESE PROJECT MANAGER
                   M. Winslow
ESE DATA MANAGEMENT      ESE QA COORDINATION
        V. O'Brien                 P. Pisigan
METHOD 5    ESE TASK MANAGEMENT    METHOD 7
       B. Weichert                G. Jackson
   HPLC	ESE ANALYSIS	GC
     W. Keeran                        J. Yen

-------
                                                                            Section No. 5
                                                                            Revision No. 2
                                                                            Date: July 29, 1990
                                                                            Page 1 of 2
5.    QUALITY ASSURANCE OBJECTIVES FOR MEASUREMENT DATA

     The "estimated detection limit" (EDL) for Method 5 will be determined as follows:

     1.    Determine the concentration of each analyte which, when injected, yields a 5:1
           signal-to-noise ratio, measured as the ratio of the center chromatographic peak
           height to the edge-to-edge height of the noise envelope.

     2.    Spike eight HPLC-grade water samples at the concentration determined in 1. and
           analyze together.

     3.    Calculate the Estimated Detection Limit (EDL) by multiplying the standard deviation
           of the concentrations from 2.  by the Student t-value 2.998 (7 degrees of freedom,
           alpha = 0.99, one-sided).

     4.    The EDL is the greater of the values calculated in 1. and 3.

     5.    The EDL shall be evaluated by the Technical Monitor.

     6.    Target EDLs shall be supplied to ESE by the Technical Monitor.

     7.    Analyze the eight sample aliquots of 2. on the confirmatory column and calculate
           EDLs as in 1 and 3.  The resulting EDLs must roughly equal those of the primary
           column.

     8.    GC/MS analyses is not part of Method 5 confirmation.

     9.    Minimum Reporting Levels  (MRL) are 3 times the EDL for each  analyte.

     10.   Report as  an occurrence (code -111) any peak matching an analyte  retention time
           at responses between one-half the MRL  and the MRL.  Such frequent occurrences
           may  lead to confirmations and/or adjustment in the reporting limit. Frequent
           occurrences of non-analyte peaks will be reported to the Technical Monitor.  Criteria
           for reporting these non-analyte peaks will be supplied by the EPA (Appendix B).

     11.   The lowest standard solution concentration will approximately equal the MRL for  all
           analytes.

     12.   Initial demonstration of method performance will be accomplished by the analysis of
           20 spiked  samples at 10 times the MRL over a four day period  (five spiked samples
           per day).  Mean recovery, variance and the  standard deviation  will be calculated.
           Relative standard  deviation (RSD) will be used to  evaluate the precision of the
           method  using the acceptance criteria provided by the EPA.

     13.   The criteria values will be reviewed by the Technical Monitor.
     Samples having any analyte concentration above that analyte's reporting limit will be confirmed

on a different HPLC  column (see Section 9).
     Performance evaluation samples  will be analyzed quarterly  as part of the overall NPS QA
program. The Technical Monitor will work with ESE to resolve any  problems discovered as a result of

these analyses.

-------
             f.--rvrrr  r-r-.    .
                                                                         Date: "\July" 29, 1
                                                                         Page 2 of 2
     Results from all of the above procedures in this Section will be reported to the Technical Monitor
for approval. Data from the Initial  Demonstration of Method Performance is appended to this Plan.
     During the survey, EPA will conduct a time-storage study and will provide extra field samples
(10% of sites) for ESE to assess analyte recoveries from a variety of matrices. Each sample will be
spiked with a stock solution in methanol to yield sample concentrations of  2, 10 or 20 times the
reporting limit for each analyte (the surrogate will also be added) (Appendix C).  This sample will be
analyzed and the data reported as a percent recovery. This data is not for laboratory control, and the
analysis will be judged to have failed only if the extra sample's surrogate recovery fails the criterion for
a regular sample.

-------
                                                                            Section No 6
                                                                            Revision No. 2
                                                                            Date:  July 29, 1990
                                                                            Page 1 of 4
6.   SAMPLING PROCEDURES
     All samples for Method 5 will be received in 250 ml_ bottles.  Samples for Method 5 will include
pH 3 buffer and will be shipped iced for overnight delivery.  Every set received from a field site will
include the following: primary samples and one backup sample for each primary sample. Some sets
will include time-storage samples. Some sets (about 10%) will include lab-spike samples (ESE matrix-
spikes).
     Analysis types and frequency are described on the next page.

-------
• r v , , rn.-r
The analysis types for Method 5
Primary Analysis:
Method Blank
Calibration Standards
Field Sample
Lab Spike (ESE Matrix Spike)
Day 0 Time-Storage Sample
Day 14 Time-Storage Sample
Performance Evaluation Sample
Backup Sample
Confirmational Analysis:
Method Blank
Calibration Standards
Field Sample
f*. V - *> "•
are:

1 per set
daily
1 per ESE
sample no.
10%
10%
10%
quarterly
(a)
1 per set
daily (when conf.
is performed)
(b)
— ^^^ ,_ . Section No. 6 . f -
x r» • f 1 ' Revisidn No. 2 ' ! I *
"Date: July 29, 1990
Page 2 of 4


Not chargeable
Not chargeable
Chargeable
Chargeable
Chargeable
Chargeable
Chargeable
(a)
Not chargeable ' — "- ~
Not chargeable
Chargeable
(a)    Analyzed when results of initial analysis fails (not chargeable) or at request of Technical
      Monitor (chargeable).

(b)    Analyzed when results of primary analysis are above min. reporting limit.

-------
                                                                      Section No. 6
                                                                      Revision No. 2  •
                                                                     "Date:  July'SS, 1990
                                                                      Page 3 of 4
Each sample's shipping label will be of the form below.
                       NATIONAL PESTICIDE SURVEY - NFS

                         PD-0415-4-5-6     DATE 2-11 -88
                            ESE - #5 - T/S0
                           SAMPLER NAME      	
The sample bottle number (PD-0415-4-5-6 in the above example) is constructed as follows:
PD        Pesticide Survey Domestic Well
PC        Pesticide Survey Community Well
PR        Pesticide Survey Resampled Site
PB        Pesticide Survey Performance Evaluation Sample
0415    •   is the site number                                               —
4          is the ESE lab number
5          is the carbamates method number
6          is the bottle number within this site number
The analysis type codes are given below:
FS        Field Sample
FD        Field Duplicate
BU        Backup Sample
LS        Lab Spike (ESE Matrix Spike;  "LS" may be followed by a numeric digit which
T/S
T/S0
T/S
   14
                indicates spike level)
                Time Storage (t = 14 days)
                Time Storage Duplicate (t = 0 days)
                Time Storage Duplicate (t = 14 days)
     The sampling contractor will supply a copy of the field sample tracking sheet used for NFS
samples.
     The spiking solution is prepared fresh daily in chloroacetic buffered water and is prepared from
a primary acetonitrile stock.  The lab spike is prepared by adding the appropriate volume of the
spiking stock (AO = 200 uL; A1  = 500 uL; A2 = 1000 uL) to 250 uL of 2.5 M chloroacetic buffer and
enough HPLC water to yield  a final volume of 25.0 mL Then 10 mL of this solution is filtered through
a 0.45 uM Acrodisc filter and spiked with 100 uL of the internal standard. All lab spikes are analyzed
on the same day that they are prepared.

-------
                                                                           "Date: July'2§, 1390
                                                                            Page 4 of 4
     In addition, the time storage samples are prepared in exactly the same manner as the above lab
spikes.  The Day 14 time storage samples are then transferred to a 40 ml amber VGA bottle and
stored in the freezer. On the day of analysis, the sample is thawed, filtered, and spiked with the
internal standard.
     Using the above spiking schemes, the solutions should correspond to: AO = 2 x MRL; A1 = 5 x
MRL; A2 = 10 x MRL

-------
f  ' I  }
Section No. 7
Revision. No 2-
"Date: July'2§,
Page 1 of 1
                                                                                             if"'
7.   SAMPLE CUSTODY
     The sampling contractor (ICF) will supply information about sample shipments and the protocol
ESE will follow to notify the sampling contractor about sample receipt and any problems associated
with samples received (Appendix D). The project Technical Monitor will also be notified concerning
problems with the receipt of samples from the sampling contractor (ie. no ice, incorrect pH,  etc).
     Holding times for samples to be analyzed by Method 5 are: 14 days maximum holding time for
samples, counting day of sampling as Day 0.  Upon receipt of samples at the laboratory, a portion of
the sample will be removed to facilitate freezing and for pH measurement using disposable pH test
strips.  Samples will be stored frozen in the dark.  Sample disposal will be handled in accordance with
Florida regulation.
     A series of time storage samples will be collected and analyzed during the NPS study to
determine the stability of the compounds in aqueous solution and in the extraction solvent.  These
samples will be collected at a frequency of 10% over the  period of the study.
     Four extra aliquots of the samples for the time storage will be collected.  Two of the four
replicate-aliquots will be spiked and analyzed within 4 days of spiking and prior to the 14 day .holding-
time. The remaining two duplicates will be spiked at the same time as the first two duplicates, but will
be allowed to sit  14 +_ 4 days before analysis.
     Each sample will be spiked with only one mix at the 10X MRL level.  Results of the time storage
samples will be reported to EPA along with the corresponding regular sample.  For data reporting
purposes, Day-0  samples (spiked and analyzed within 4 days of spiking) will be referred to as Day-0
Time Storage (DTS).  These samples serve a dual role, as lab spikes  and DTS.  Day-14 time storage
samples (spiked  and  held for 14 +_ 4 days before analysis) will be referred to as Holding Time
Samples (HTS).  All of the holding time samples will be analyzed in duplicate.
     When samples  are received at the ESE Receiving Station, 14420 Newberry Road, Gainesville,
FL, the receiving  employee will deliver the samples to Vince Prem-Das at the sample check-in station,
the Hendrickson  building.   Mr. Prem-Das will then mark the NPS sample tracking form with the
required information, and submit the logsheets to Data Management.  A Data Management employee
will enter the sample information to ESE's database, a process which automatically generates Sample
Arrival Notices for the Method 5 Task Manager. The GC/HPLC department schedules the analyses,
performs them within  the 14-day holding times, and enters the analytical data into the ESE database.
     All ESE coldrooms, refrigerators, and freezers holding samples  for chemical analysis are
monitored daily by ESE personnel, and records are kept  daily.

-------
                   r  n
r
Section No 8
Revisioh No. 2'
"Date: July'2!,
Page 1 of 1
8.    CALIBRATION PROCEDURES AND FREQUENCY
      The instrumental analysis for Method 5 will be performed on a HPLC instrument including an
Shimadzu Gradient HPLC System: 2 LC-6A pumps, SCL-6A system controller, SIL-6A autoinjector, a
Kratos post-column reactor, a Shimadzu RF535 fluorescence detector, and a PE/Nelson
chromatography data acquisition system.  Primary analytical separations will be performed on Waters
Nova Pak ODS columns (25 cm length, 4.6 mm ID, 5 micron particle size) or equivalent, and
secondary analytical separations will be performed on Supelcosil LC-1 columns (25 cm length, 4.6 mm
ID, 5 micron particle size).  The mobile phase gradient program will be that described in Method 5
which is appended to this Plan.
      ESE analysts will prepare standard solutions in water/chloroacetic acid buffer from dilutions of
EPA concentrate solutions at the onset of each analytical set and will keep detailed records of the
means used to prepare them on each occasion. Spiking concentrates will be generated by separate
dilutions from those used to generate standard solutions.  Records of the generation of standards will
be kept by the analyst in his permanent notebook. Each set's analytical records will refer  to the
standard" solutions actually-used in that day's analysis.  New standard dilutions will be checke4-to
insure that the QC criteria, ± 20% of initial calibration, is met.
      ESE plans to calibrate the instrument each day a set is  analyzed using a calibration standard
concentration chosen, on a rotating basis, from the standard curve.  The lowest calibration standard
will be at the MRL The standard relative response must agree within +_ 20% of the previous set's
calibration relative response.
      For  second column confirmation, a single calibration standard will be used for quantitation.  The
standard  will contain  analytes  at concentrations near that detected during primary analysis. Results of
second column confirmation will be within +_ 25%  of the primary analysis result. If this criteria is not
met the project Technical Monitor will be advised.

-------
                  r    n.
 Section No. 9
 Revision No. 2-  •
"Date: July 25, 1990
 Page 1 of 1
9.    ANALYTICAL PROCEDURES
     Samples will be thawed and brought to room temperature in a warm water bath prior to
analysis.  The primary analysis begins with the addition of internal standard to an aliquot of water
sample. Gradient elution down the reverse-phase HPLC column separates the analytes and other
components, and analytes are detected as follows: the analytes are hydrolyzed with 0.05 N sodium
hydroxide at 95 °C, the resulting methylamine is reacted with ortho-phthalaldehyde (OPA) and 2-
mercaptoethanol, and the resulting highly fluorescent derivative is detected in the flow cell of a
fluorescence detector.  In confirmatory analysis, the sample aliquot is eluted down a somewhat
different column, detection in the same manner verifies or rules out any positive results from the
primary analysis.
     Primary and confirmatory analyses are performed on an Shimadzu  high-pressure liquid
chromatograph equipped with a Kratos post-column  reaction cell, a Shimadzu fluorescence detector
(excitation wavelength  = 235nm, emission wavelength =  435nm), and a PE/Nelson chromatography
data acquisition system.  All data will be transferred to the Chemistry Division database running on 65
AT-class-computers served by a 3-drive, 1600-Mbyte Novell network housed in the Chemistry Division.--
     ESE plans to include no more than 24 field samples in a set (16 hour run), and generally fewer.
     ESE plans no significant differences from Method 5, revision dated October 27, 1987 (attached)
as provided to ESE in glassware cleaning, reagents,  or data reduction. Differences in equipment  are
listed above in this Section.  In the analytical procedure, ESE plans, to use Waters Nova Pak ODS
columns for the primary analysis and Supelcosil LC-1 columns for the confirmatory analysis.
Any deviations from these procedures or QC requirements after the Plan  is approved will only be  used
if approved by the EPA Technical Monitor in advance. Changes will be documented, signed  by the
Technical Monitor and appended to the Plan.

-------
                  r    n /
 Section No. 10
 Revision No. 2  •
"Date:  July 2:9, 1S90
 Page 1 of 1
'.f
10.   DATA REDUCTION, VALIDATION REPORTING
      Calibration standard concentrations and raw instrument responses for calibration standards and
sample aliquots are entered into the ESE Chemistry Division's database.  ESE data management will
support new in-house sample type codes that the numerous EPA sample types require. The analyst
will generate ESE set reports and will verify correct entry and QC data compliance with project criteria.
When Data Management receives a signed copy of the set report, they will generate a data file on
floppy disk in the NPS format.  The project manager will then examine a listing or summary of the file
and approve delivery of the floppy disk to EPA.
      All data for a set of samples (including QC and confirmatory data) will be reported to the EPA
no later than two months from the date of sample collection.
      The NPS file format is appended to this Plan (Appendix E,  F and  G).
      Fast Track Reporting (immediate telephone call to the Technical monitor) will be needed for
confirmed positive sample concentrations of selected EPA analytes above health advisory  levels and
situations when results from confirmation columns do not agree with results from primary columns
within criteria set by  EPA (+ 25%) (Appendix H.).                                         —
      Storage of laboratory data will be by ESE  standard procedures.  Standards and reagent
preparation data will reside in the analyst's permanent laboratory notebook. Chromatograms,
calibration data, and corrective action  records, etc. will reside in an analysis set folder which will be
stored permanently in ESE's central filing system.  ESE set folders will reside in archival files in ESE's
central filing system after all samples have been analyzed.

-------
i ^
                                                                             Section No. 1 1
                                                                             Revision No. 2  -
                                                                             *        * -Y   j .
                                                                             Date: July 29, 1930
                                                                             Page 1 of 3
                                                                    '.f  •,
11.   INTERNAL QUALITY CONTROL CHECKS
      No lab control spikes are required for Method 5.  Analytical quality will be controlled largely by
five means:
           Each set will include a chromatogram of a standard, selected on a rotating basis
           from the calibration standards.  The response factors for the standard must be
           within +_ 20 percent (%) of the initial calibration.  If the response factors fall outside
           the +. 20% criteria, the standard will be  reanalyzed, if the response factor is stilt
           outside the criteria a new initial calibration will be performed.
           Each set's method blank must have analyte responses less than half the minimum
           reporting limit.
           Internal standard response (peak area) for each  sample must be within 20% of the
           internal standard response from that day's calibration standard.
           The instrument performance sample will meet the criteria described in Method 5,
           Section 10.8., Table 10.
           If the relative responses of the analytes  in each set's standard chromatograms
           appear to the analyst to have changed significantly (+_ 20%) from those of the.
           previous set, lie will make new standard solutions from the EPA concentrates. The
           stability of Method 5 analytes suggests that this may be a very  unusual occurrence.
      •     Any analyte detected above the MRL in  the primary analysis will be re-analyzed on
           the confirmation column.  Results of the confirmation analysis must agree within +_
           25% of the primary analysis.  Any deviation from  this criteria will be reported to the
           project Technical  Monitor.
      No laboratory control standards will be analyzed, and surrogates are  not used. The internal
standard, 4-bromo-3,5-dimethylphenyl-N-methyl-carbamate (BDMC), supplied by  USEPA, will be  added
to every sample, standard, and method blank before analysis. Internal standard recovery must be
evaluated for every sample by determining whether  the measured  peak area for the internal standard
in any sample or method  blank deviates by more than 20% from the peak area for the internal
standard in that day's calibration standard.  In case of such disagreement,  reinject the sample.  If the
second analysis still shows disagreement, add internal standard to a second aliquot of the sample
and reanalyze.
      However, if more than one sample shows disagreement with the internal standard response of
the calibration standard, reanalyze the instrument performance standard, the calibration standard, and
the samples which showed the disagreement.  If the disagreement still persists, contact the Technical
Monitor.
      A method blank is analyzed daily. If any analyte is present at half or  more of the MRL, the test
is considered to have failed, and a new method blank will be generated and analyzed. -A second
consecutive method blank failure will require in-lab corrective action and reanalysis starting with
instrument performance standard and calibration standards.

-------
—
 fa '
 «- fc. 4 *•
 Section No. 1 1
 Revision No. 2  •
•       '«,   t.
 Date:  July 29, 1990
 Page 2 of 3
                                                                                              i  f  •  '
                                                                                               * F   9
      Each day that NPS Method 5 samples are analyzed, an instrument performance sample will be
analyzed and the instrument sensitivity (signal to noise ratio), chromatographic performance (peak
symmetry factor and peak gaussian factor), and the column  performance (resolution) will be monitored
as described in Method 5.
      During confirmational  analyses, a method blank and standard will be analyzed prior to analysis
of samples.  All quality control criteria which apply to the primary analysis, with the exception of the
instrument performance standard, will also apply to the confirmational analysis. All confirmation data
and QC data will be stored with the set in which the field sample number resides.
      ESE plans to have in place an NPS option in the ESE  set report program. This option will check
that all QC is performed, just as is done for all batches currently reported under ESE's standard QC
program.
      Exhibit 11-1 presents ESE's Internal Quality Control Checklist.
      Since  (1) ESE will demonstrate its capability to accurately perform Method 5, (2) the
instrumental analysis will most often  be performed within only hours of first opening the sample bottle,
(3) EPA-will be-sending check samples, (4) the internal standard response  on each sampte will-be
controlled, (5) the instrument performance sample will be analyzed in each set, and (6) the  above QC
checks will be performed on each set,  ESE considers the above program sufficient to assure the
quality of the data and does not propose additional QC  checks.
      Quality-Control Corrective Actions  - Failure to meet QC  criteria specified in this QA plan.
Actions consist of two kinds: those resolved within each analytical department by reanalysis, etc.; and
those resolved outside the department. All corrective actions will be reported to EPA's Technical
Monitor.

-------
                                  _                    _              Section No. 11
             { V    '     r"   V    ^   *     ., ^  f*±  t '.   •  Revision No. 2  .    'if-'
                                                       "'   *    "    "Date:  July &, 1990
                                                                      Page 3 of 3
                               EXHIBIT 11-1

                INTERNAL QUALITY CONTROL CHECKLIST
                 U. S ENVIRONMENTAL PROTECTION AGENCY
                        NATIONAL PESTICIDE SURVEY
                                 METHOD 5
                    INTERNAL QUALITY CONTROL CHECKS
EPA SET No.	 ESE BATCH No._

DATE ANALYZED  	           ANALYST
                                                            YES         NO

      CHROMATOGRAM OF DAILY STANDARD INCLUDED      	      	

      METHOD BLANK (< 0.5 Minimum Reporting Limit)            	      	
3.     INTERNAL STANDARD RESPONSE (Peak Area)
      (Each Sample +. 20% of Daily Standard)

4.     INSTRUMENT PERFORMANCE SAMPLE

      A     Sensitivity (Signal/Noise > 3.0)

      B     Chromatogivipluc Performance

            (1)    Peak Symmetry Factor (PSF) (0.9 1.0)

5.     DAILY STANDARD ANALYTE RESPONSE
      (Each Analyte +_ 20% of Previous Daily Std.)

6     CONFIRMATION (± 25 % of Primary Analysis)


COMMENTS:	   	    	      	
QUALITY ASSURANCE COORDINATOR	  DATE_

PROJECT MANAGER	    	       	  DATE

-------
                   r   n
 Section No 12
 Revision No. 2
"Date:  July 2*9,
 Page 1 of 1
12.   PERFORMANCE AND SYSTEM AUDITS

      ESE's Quality Assurance Division, independent of the Chemistry Division and reporting directly

to the ESE president, will perform audits of the following types:

           Observation of the analyst analyzing samples during the initial demonstration of
           capabilities and approximately quarterly thereafter.  This audit will include verification
           that no significant changes have occurred  in procedure, instrumentation, analytical
           environment, or in sample and reagent storage and labelling. The auditor will select
           items for audit from this QA Plan at his discretion and generally without warning to
           the analyst.  Retrievable errors (those affecting no data  yet sent to EPA), will be
           corrected immediately and a means of assuring its long-term rectification
           established.  Irretrievable errors will prompt written notice to the EPA Technical
           Monitor.

           Examination of a data set, especially QC and instrument performance parameters.
           This will be performed in detail during the initial demonstration of capabilities and
           approximately quarterly thereafter.  The latter examinations will be performed on
           randomly selected data sets.  The auditor may question the analyst at his discretion.
           However, the analyst may postpone for up to 10 working days a prolonged
           conference with the auditor if (1) the analyst or his department manager considers
           the. current backlog of samples, especially  those near holding times, to be too great
           to allow immediate consideration of the auditor's questions AND (2) if the auditor's"
           questions do not concern data due within a few working days.

      In addition to the above in-house audits, EPA will perform audits of the data and systems as the

pertain to NPS Method  5.  EPA audits will be scheduled approximately every six months during the

conduct of Method 5 analyses.  ESE will cooperate with the EPA auditor to the extent that the audit

does not interfere with the laboratories ability to continue with routine operation.

-------
                       n
Section No. 13
Revision No. 2
"Date: July'29, 1990
Page 1 of 1
13.  PREVENTIVE MAINTENANCE

     The instrumentation ESE specifies for the performance of Method 5 requires little preventive

maintenance.
     •     Check valves will be replaced as needed.

     •     Pumps are tested for flow rate accuracy when, (1) the analyte retention times
           change between analytical sets or (3) the pumps are changed or serviced.

           Analytical columns will be protected by use of 3 to 5 cm length pellicular guard
           columns.  Guard columns will be changed when deterioration of the analytical
           column chromatography (e.g. poor peak shape, loss of resolution, etc.) becomes
           evident.

     •     Spare columns,  guard-column packing material, instrument cables, and some PC
           boards are kept to minimize instrument downtime.

-------
                                                                             Section No  14
                                                                             Revise/No 2-
                                                                            -Date: ju|y'j^
                                                                             Page 1 of 1
14.   SPECIFIC PROCEDURES for ASSESSING MEASUREMENT SYSTEM DATA
      Relative response of each analyte is calculated as
                RRa =  Aa/Ais
      where:     Aa  =  peak area of the analyte, and
                Ais =  peak area of the internal standard from the same chromatogram.
      During initial demonstration of capabilities, the mean recovery is calculated as the sum of
recoveries divided by the number of recoveries included in the sum.  The variance is calculated as the
sum of the squares of differences between each recovery  and the mean recovery, divided by the
number of recoveries less one. The standard deviation is  the square root of the variance. The relative
standard deviation (RSD) is calculated to evaluate precision performance. RSD is calculated as:
           RSD = (standard deviation of recoveries/mean recovery) X 100
      Each sample's  chromatogram's internal standard response (peak area) must be within a range
of 0.8 to 1.2 times the mean internal standard response for that day's calibration curve
chromatograms.
      Asrnentroned in Section 11, a single standard will be analyzed daily as a check of instrument  -
sensitivity.
      Responses of peaks corresponding to the analytes of interest found in the samples (peak area
>1/2 MRL) will be subtracted from the responses of analytes found in the spiked samples to correct
for concentrations of  analytes present in the sample prior to spiking.  Found concentrations  in the
spiked samples will be divided by the "true" value and multiplied by 100 percent to calculate recovery.
      An instrument quality control standard will be analyzed with each analytical set and the criteria
listed in Method 5, Table 10 will be used to assess  instrument performance. Due to coelution of
methiocarb and the internal standard on the confirmation,  the resolution test can not be performed on
this column.

-------
                   r    n
r
 Section No. 15
 Revision No. 2-  •
"Date: July'25, 1990
 Page 1  of 2
15.   CORRECTIVE ACTION

      In general, bench-level corrective actions fall into two categories each with differing required
action.
           Short-Term Action  -  Major and minor problems which can be corrected
           immediately.  Examples include failure to date or sign a field form and date entry
           errors.  Generally, the analyst or other employee committing the error can simply
           correct it, and the record of this corrective action will directly reflect the  error and its
           resolution.

           Long-Term Corrective Action   - Minor and major problems which require a series of
           actions to resolve the problem.  Examples include a discovery that part  of the
           analytical or data-handling procedures were not being followed correctly.  The
           actions to be taken are coordinated by the QA Supervisor or his designate, and a
           QA corrective action and routing form (Figure 15-1) is used to track the  action.
           These corrective actions and their resolutions will be included  in monthly reports to
           EPA.

-------
              n
                Section No. 15
                Revision No. 2
               "Date:  July 29, 1§90
                Page 2 o< 2
                               EXHIBIT 15-1
                QUALITY ASSURANCE CORRECTIVE ACTION
                           AND REQUEST FORM
           QUALITY ASSURANCE CORRECTIVE ACTION REQUEST
                           AND ROUTING FORM
1.  Identification of a Problem:
   Originator:
   Nature of Problem:
2.  Determination of Required Action:
   Responsibility Assigned to:
   Recommended Action:
3.  Implementation of Required Action:
   Responsibility Assigned to:
4.  Assuring Effectiveness of Action:
   Responsibility Assigned to :
   Procedure to assure Effectiveness:
Date:
Due Date:
Due Date:
Due Date:

-------
                                                              __              Section No. 16
                          \  "*  (**  (P*   V,    *•  **    \ ^^^  f\ t     '   Revision No. 2  .
                               ''         	'    "' '*    -   Date: July'29, 1990
                                                                             Page 1 of 1
16.   QUALITY ASSURANCE REPORTS TO MANAGEMENT

      QA activities are reported to management by the QA Coordinator in three ways:
      1.    Verbal notification of significant QA deficiencies immediately upon discovering the
           problem,

      2.    Written interim QA reports, and

      3.    Written final QA reports.

      A final QA report will be prepared for this project.  Interim reports will be prepared at the request

of ESE management, the  Project Manager, or the Contracting Officer.

      The contents of interim and final QA reports will be similar except final reports will include
summaries of the  interim reports.

      The following items will be addressed in the reports:

      1.    An assessment of the precision and accuracy data associated with sample data
           generated during the report period.

      2.    Results of all  QA audits performed during the report period.

      3.    Results of the QA data validations performed during the report period.
      ESE will send monthly reports to the EPA Technical Monitor. These reports will be in the

following format:

           Summary of progress  -     samples received and samples analyzed, but not
                                      validated and status of  data processing for analyzed
                                      sets of samples and numbers for sets of data sent to
                                      the EPA

      •     Reports on standards  -     new dilutions and results of checks before using

           Summary list of bench -     level corrective actions  (as in Section 15 of this
                                      Plan).

           Identification of problems  about any phase of the project.

      •     Copies of representative and, if applicable, unusual chromatograms (as
           requested by Technical Monitor).

           Information requested by  the Technical Monitor because of specific
           methodology or problems encountered.

           Changes in personnel.

           Comments

-------
                                                                             Section No. 17
                                                                            .Revision NJD 2
                                                                            *Date: July 29,
                                                                             Page 1 of 2
17.  ARCHIVAL OF RAW DATA
     In order to assure the continued availability of all documentation necessary to defend NFS
Method 5 analytical results, the documentation for each analytical set will be contained in a separate
orange file folder and will be comprised of the following:
     •     HPLC set documentation checklist
     •     Method 5 internal quality control checklist
     •     A hard copy of the NPS formatted results
           A hard copy of ESE's formatted results
     •     All chromatograms and quantitation reports for the following:
                 Field samples
                 Blanks
                 Calibration standards
                 Instrument performance standard
                 Time-storage samples
           --  '   Lab spike samples
           Chromatographic Analysis logsheets
           Instrument logsheet(s)
     •     Internal Standard Recovery summary
           Copies of NPS sample tracking forms
     •     Copies of analyst notebook pages
           Standard Curve data sheet
           Instrument Configuration sheet
     All set file folders will be stored in banker's boxes.  Each box will be labeled with number, a
description of its contents (NPS Method 5 set files), a listing of set numbers, the date placed in
storage, and the date to be destroyed.
     A separate banker's box will contain the following, and be labeled  as described above:
     •     A copy of the QA Project Plan
     •     Monthly  reports to the Technical Monitor
           Original NPS sample tracking forms with original Fed Ex airbills
           Correspondence to and from  EPA
     •     Laboratory data from initial demonstration of capabilities and
           demonstration of accuracy and precision
           Miscellaneous documentation such as audit reports, internal memos, temperature
           logs, etc.
     •     Resumes of NPS participants

-------
M
                                                                             Section No. 17
                                                                             Rev'sion No. 2
                                                                             "Date: July 2fe, 1990
                                                                             Page 2 of 2
     All banker's boxes containing the NFS files will be stored for a period of seven years in the
Dead File Storage room located in the Maintenance building at ESE's Gainesville location. This room
is kept locked at all times and access is limited to its custodian, Virginia O'Brien, Manager of the
Information Services Department of ESE's Gainesville laboratory.
     A Dead File Storage logsheet listing the same information that is written on bankers boxes will
be kept with the storage room custodian  and a copy will be sent to the EPA Technical Monitor.  In
addition, a list of all samples analyzed will be prepared, with cross references to the NFS set number,
the ESE batch number, and the ESE internal sample number.
     Appendix J provides a copy of ESE's standard operating procedure for batch storage of data.
Any changes to the archival procedures described above will be communicated in writing to the
Technical Monitor.

-------
                                                      Appendix A
                                                      Revision No. 2    ",  I
                                                      "Date: July'25, 1990   *'
                                                      Page 1 of 27
                        APPENDIX A

METHOD 5. MEASUREMENT OF N-METHYLCARBAMOYLOXIMES AND
 N-METHYLCARBAMATES IN GROUND WATER BY DIRECT AQUEOUS
     INJECTION HPLC WITH POST COLUMN DERIVATIZATION

-------
            Method 5.   Measurement of N-Methylcarbamoyloximes and
             N-Methylcarbamates  in Ground Water by Direct Aqueous
                Injection HPLC with Post Column Derivatization
1.        SCOPE AND APPLICATION

         1.1    This is a high performance liquid chromatographic (HPLC)
                method applicable to the determinations of certain
                N-methylcarbamoyloximes and N-methylcarbamates in ground
                water.   Analytes that can be determined using this method are
                listed in Table 1.

         1.2    This method has been validated in a single laboratory.
                Estimated detection limits (EDLs) have been determined and
                are listed in Table 3.   Observed detection limits may vary
                between ground waters,  depending upon the nature of
                interferences in the sample matrix and the specific
                instrumentation used.

         1.3    This method is restricted to use by or under the supervision
                of analysts experienced in the use of liquid chromatography
                and in the interpretation of liquid chromatograms.   Each
                analyst must demonstrate the ability to generate acceptable
                results with this method using the procedure described in
                Section 10.2.

         1.4    When this method is used to analyze unfamiliar samples for
                any or all of the analytes above, analyte identifications
                must be confirmed by at least one additional qualitative
                technique.

2.        SUMMARY OF METHOD

         2.1    The water sample is filtered and a 400-uL aliquot is injected
                into a reverse phase HPLC column.  Separation of the analytes
                is achieved using gradient elution chromatography.   After
                elution from the HPLC column, the analytes are hydrolyzed
                with 0.05 N sodium hydroxide (NaOH) at 95°C.  The methyl
                amine formed during hydrolysis is reacted with
                o-phthalaldehyde (OPA)  and 2-mercaptoethanol to form a highly
                fluorescent derivative which is detected by a fluorescence
                detector.l

3.        DEFINITIONS

         Artificial ground water -- an aqueous matrix designed to mimic
         characteristics of a real ground water sample.  Artificial ground
         waters should be reproducible  for validations performed in other
         laboratories.

-------
       r  •• n
Calibration standard -- a known amount of a pure analyte,  dissolved
in an organic solvent,  that is analyzed under the same procedures
and conditions that are used to analyze samples containing that
analyte.  '

Estimated detection limit (EDL) -- the minimum concentration of a
substance that can be measured and reported with confidence that the
analyte concentration is greater than zero as determined from the
analysis of a sample in a given matrix containing the analyte.

Field duplicates -- two samples, collected at the same site, that
are treated exactly the same throughout field and laboratory
analytical procedures.   Analysis of field duplicates provides a
measure of the precision associated with sample collection,
preservation and storage, as well as with laboratory procedures.

Instrument quality control (QC) standard --an aqueous solution
containing specified concentrations of specified analytes.   The
instrument QC standard is analyzed each working day prior  to the
analysis of sample extracts and calibration standards.  The
performing laboratory uses this solution to demonstrate acceptable
instrument performance in the areas of sensitivity, column
performance, and chromatographic performance.

Internal standard - - a pure compound added to a sample in  a known
amount and used to calibrate concentration measurements of other
analytes that are sample components.   The internal standard must be
a compound that is not a sample component.

Laboratory control standard --a solution of analytes prepared in
the laboratory by dissolving known amounts of pure analytes in a
known   amount of reagent water.  In this method, the laboratory
control           standard is prepared by adding appropriate volumes
of the appropriate standard solution to buffered reagent water.

Laboratory reagent blank -- an aliquot of buffered reagent water,
filtered, and analyzed as if it were a sample.

Quality control check sample --a water-soluble solution containing
known concentrations of analytes prepared by a laboratory  other than
the laboratory performing the analysis.  The performing laboratory
uses this solution to demonstrate that it can obtain acceptable
identifications and measurements with a method.   A small measured
volume of the solution is added to a known volume of buffered
reagent water and analyzed with procedures identical to those used
for samples.  True values of analytes are known by the analyst.

Stock standard solution --a concentrated solution containing a
certi-fied standard that is a method analyte, or a concentrated
solution of an analyte  prepared in the laboratory with an  assayed
reference compound.

-------
                                                                              ,  ,
                                                                               • *
4.        INTERFERENCES
         4.1    Method interferences may be caused by contaminants in
                solvents, reagents, glassware and other sample processing
                apparatus that lead to discrete artifacts or elevated
                baselines in liquid chromatograms.   Specific sources of
                contamination have not been identified.  All reagents and
                apparatus must be routinely demon-strated to be free from
                interferences under the conditions of the analysis by running
                laboratory reagent blanks as described in Section 10.6.1.

                4.1.1     Glassware must be  scrupulously cleaned.2  Clean all
                          glass-ware as soon as  possible after  use  by
                          thoroughly rinsing with the  last solvent  used in
                          it.   Follow by washing with  hot water and detergent
                          and thorough rinsing with  tap and reagent water.
                          Drain dry,  and heat in an  oven or muffle  furnace at
                          450°C for 1 hour.   Do  not  heat volumetric ware.
                          Thermally stable materials might not  be  eliminated
                          by this treatment.   Thorough rinsing  with acetone
                          may be substituted for the heating.   After  drying
                          and cooling,  seal  and  store  glassware in  a  clean
                          environment to prevent any accumulation  of  dust or
                          other contaminants.   Store inverted or capped~with
                          aluminum foil.

                4.1.2     The use of high purity reagents and solvents  helps
                          to minimize interference problems.  Purification of
                          solvents by distillation  in  all-glass systems may be
                          required.

         4.2    Interfering contamination may occur when a sample containing
                low concentrations of analytes  is analyzed immediately
                following a sample containing relatively high concentrations
                of analytes.  A preventive technique is between-sample
                rinsing  of the sample syringe and filter holder with two
                portions of reagent water.  After analysis of a sample
                containing high concentrations  of analytes, one or more
                laboratory reagent blanks should be analyzed.

         4.3    Matrix interference may be caused by contaminants that are
                present  in the sample.  The extent of matrix interference
                will vary considerably from source to source, depending  upon
                the ground water  sampled.  Positive identifications must be
                confirmed using  the confirmation column specified in Table  2.
5.  SAFETY
         5.1    The  toxicity or carcinogenicity of each reagent used in this
                method has not been precisely defined; however, each chemical
                compound must be  treated as a potential health hazard.  From
                this viewpoint, exposure to these chemicals must be reduced
                to the lowest possible  level by whatever means available.
                The  laboratory is responsible for maintaining a current

-------
                awareness file of OSHA regulations regarding the safe
                handling of the chemicals specified in this method.  A
                reference file of material data handling sheets should also
                be made available to all personnel involved in the chemical
                analysis.  Additional references to laboratory safety are
                available and have been identified3"5 for the information of
                the analyst.
6.
APPARATUS AND EQUIPMENT (All specifications are suggested.
numbers are included for illustration only.)
Catalog
         6.1
         6.2
         6.3
         6.4
       SAMPLING EQUIPMENT
       6.1.1     Grab sample bottle --  60-mL screw cap vials (Pierce
                 No.  13075 or equivalent)  and caps equipped with a
                 PTFE-faced silicone septa (Pierce No. 12722 or
                 equivalent).  Prior to use,  wash vials and septa
                 with detergent and rinse  with tap and distilled
                 water.   Allow the septa to air dry at room
                 temperature, place in  a 105°C oven for 1 hour, then
                 remove  and allow to cool  in  an area known to be free
                 of organics.  Heat vials  at  400°C for 1 hour to
                 remove  organics.

       Balance -- Analytical, capable of accurately weighing to the
       nearest 0.0001 g.

       FILTRATION APPARATUS

       6.3.1     Macrofiltration -- to  filter derivatization
                 solutions and mobile phases  used in HPLC.   Recommend
                 using 47 mm filters (Millipore Type HA,  0.45 urn for
                 water and Millipore Type  FH,  0.5 urn for organics  or
                 equivalent).'

       6.3.2     Microfiltration -- to  filter samples prior to  HPLC
                 analysis.   Use 13 mm filter  holder (Millipore
                 stainless steel XX300/200  or equivalent),  and  13  mm
                 diameter 0.2 urn polyester  filters (Nuclepore 180406
                 or equivalent).

       SYRINGES AND SYRINGE VALVES

       6.4.1     Hypodermic syringe --  10-mL  glass,  with Luer-Lok
                 tip.

       6.4.2     S , _inge valve --  3-way (Hamilton HV3-3 or
                 equivalent).

       6.4.3     Syringe needle  - - 7 to 10-cm long,  17-gauge, blunt
                 tip.

       6.4.4     Micro syringes  -- various  sizes.

-------
6.5    MISCELLANEOUS

       6.5.1     Solution storage bottles --  Amber glass,  10-  to
                 15-mL capacity with TFE-fluorocarbon-lined screw
                 cap.

       6.5.2     Helium,  for degassing dissolved oxygen.

6.6    HIGH PERFORMANCE LIQUID CHROMATOGRAPH (HPLC)

       6.6.1     HPLC  system capable of injecting 200 to  400-uL
                 aliquots,  and performing binary linear gradients at
                 a constant flow rate.

       6.6.2     Primary column -- 250 mm x 4.6  mm I.D.  stainless
                 steel packed with 5 urn Altex Ultrasphere  ODS.
                 Validation data presented in this method were
                 obtained using this column.  Alternate columns may
                 be used in accordance with the  provisions described
                 in Section 10.2.

       6.6.3     Confirmation column -- 250 mm x 4.6  mm I.D.
                 stainless steel packed with 5 urn Supelco  LC-1.

       6.6.3     Detector -- Post column derivatization detector
                 composed of a post column reactor and a  fluorescence
                 detector.   This detector has proven  effective  in the
                 analysis of spiked reagent and  artificial ground
                 waters.   The post column derivatization  detector
                 (PCD) was used to generate the  validation data
                 presented in this method.   A block diagram of  the
                 PCD is shown in Figure 2.

                 6.6.3.1   Post  column reactor --  Capable  of mixing
                           reagents  into  the  mobile phase.  Reactor
                           should be  equipped with pumps to deliver
                           0.1  to 1.0 mL/min  of  each  reagent; mixing
                           tees;  two  1.0-mL delay  coils, one thermo-
                           stated at  95°C;  and constructed using  PTFE
                           tubing (Kratos URS 051  and URA  100 or
                           equivalent).

                 6.6.3.2   Fluorescence detector - - Capable of
                           excitation at  230  nm  and detection of
                           emission  energies  greater  than  418 nm.  A
                           Schoffel  Model  970 fluorescence detector
                           was  used  to generate  the validation  data
                           presented in this  method.

-------
7.        REAGENTS AND CONSUMABLE MATERIALS

         7.1    Reagent water -- reagent water used to generate the
                validation data in this method was distilled water obtained
                from the Magnetic Springs Water Co., 1801 Lone Eagle St.,
                Columbus, Ohio  43228.

         7.2    Methanol -- Distilled-in-glass quality or equivalent.

         7.3    HPLC MOBILE PHASE

                7.3.1    Water  --  HPLC  grade  (available  from Burdick and
                         Jackson).

                7.3.2    Methanol  --  HPLC grade.   Filter and degas  with
                         helium before  use.

         7.4    POST COLUMN DERIVATIZATION SOLUTIONS

                7.4.1    Sodium hydroxide,  0.05  N --  Dissolve 2.0  g of sodium
                         hydroxide  (NaOH)  in  reagent  water.   Dilute to 1.0 L
                         with reagent water.   Filter  and degas with helium
                         just before  use.

                7.4.2    2-Mercaptoethanol  (1+1)  -- Mix  10.0 mL  of
                         2-mercapto-  ethanol  and 10.0  mL of  acetonitrile.
                         Cap. Store in  hood (CAUTION  - stench).

                7.4.3    Sodium borate  (0.05  N)  --  Dissolve  19.1 g  of  sodium
                         borate (Na2B(,07.10H20) in reagent water.  Dilute to
                         1.0 L  with reagent water.  The  sodium borate  will
                         completely dissolve  at  room  temperature if  prepared
                         a day  before use.

                7.4.4    OPA reaction solution --  Dissolve 100 + 10  mg of
                         o-phthalaldehyde  (mp 55-58°C) in 10 mL  of  methanol.
                         Add to 1.0 L of  0.05 N  sodium borate.  Mix, filter,
                         and degas with helium.   Add  100 uL  of
                         2-mercaptoethanol  (1+1)  and mix.  Make up  fresh
                         solution daily.

         7.5    Monochloroacetic acid  buffer (pH3) --  Prepare by  mixing 156
                mL of 1.0 M monochloroacetic acid and  100 mL  1.0  M  potassium
                acetate.

         7.6    4-Bromo-3,5-dimethylphenyl N-methylcarbamate  (BDMC)  --  >98%
                purity, for use  as  internal standard (available from Aldrich
                Chemical  Co.).

         7.7    ARTIFICIAL  GROUND WATERS --  Two artificial ground waters were
                used to generate the validation data in this  method.   The
                first artificial ground  water  was used to mimic a hard  ground
                water,  and  the second  artificial  ground water was used  to
                mimic a ground water with moderately high organic content.

-------
       7.7.1     Hard artificial  ground water  -- The hard  artificial
                 ground water used  to  generate  the validation data in
                 this method was  Absopure Natural Artesian Spring
                 Water obtained from the Absopure Water  Company in
                 Plymouth, Michigan.

       7.7.2     Organic-contaminated  artificial ground  water --  The
                 organic-contaminated  artificial ground  water used to
                 generate  the validation data  in this  method  was
                 reagent water spiked  with  fulvic acid at  the 1 mg/L
                 concentration level.  A very well-characterized
                 fulvic acid, available from the International Humic
                 Substances Society (associated with the United
                 States Geological  Survey in Denver, Colorado),  was
                 used.

7.8    STOCK STANDARD SOLUTIONS  (1.00  ug/uL)  -- Stock  standard solu-
       tions may be purchased as  certified  solutions  or  prepared
       from pure standard materials using  the  following  procedure:

       7.8.1     Prepare stock standard solutions by accurately
                 weighing  approximately 0.0100  g of pure material.
                 Dissolve  the material in HPLC  quality methanol and
                 dilute to volume in a 10-mL volumetric  flask.
                 Larger volumes may be used at  the convenience of the
                 analyst.  If compound purity is certified at 96% or
                 greater,  the weight may be used without correction
                 to  calculate the concentration of the stock
                 standard.  Commercially prepared stock  standards may
                 be  used at any concentration if they  are  certified
                 by  the manufacturer or by  an independent  source.

       7.8.2     Transfer  the stock standard solutions into TFE-
                 fluorocarbon-sealed screw  cap  vials.  Store  at room
                 temperature and  protect from light.

       7.8.3     Stock standard solutions should be replaced  after
                 two months or sooner  if comparison with laboratory
                 control standards  indicates a  problem.

7.9    INTERNAL STANDARD  SPIKING  SOLUTION  --  Prepare  an  internal
       standard spiking solution  by accurately weighing  0.0010 g of
       pure BDMC.  Dissolve  the  BDMC in pesticide-quality methanol
       and dilute to volume  in a  10-mL volumetric  flask.  Transfer
       the internal standard spiking solution to a TFE-fluorocarbon-
       sealed scr; , .ip bottle and store  at room temperature.
       Addition of   4L of the internal standard spiking solution  to
       50 mL of sample results in a final  internal standard concen-
       tration of 10 ug/L.

7.10   INSTRUMENT QC STANDARD --  Prepare  instrument QC standard
       concentrate by adding 20  uL of the  3-hydroxycarbofuran stock
       standard solution,  1.0 mL of the aldicarb sulfoxide stock
       standard solution,  200 uL of the methiocarb stock standard

-------
                solution, and 1 mL of the internal standard spiking solution
                to a 10-mL volumetric flask.  Dilute to volume with methanol.
                Thoroughly mix concentrate.  Prepare instrument QC standard
                by placing 100 uL of the concentrate solution into a 100-mL
                volumetric flask.  Dilute to volume with buffered reagent
                water.  Thoroughly mix instrument QC standard and transfer to
                a sealed bottle.  Store at room temperature.

         SAMPLE COLLECTION.  PRESERVATION AND HANDLING

         8.1    Grab samples must be collected in glass containers.
                Conventional sampling practices8  should be followed;  however,
                the bottle must not be prerinsed with sample before
                collection.

         8.2    SAMPLE PRESERVATION/PH ADJUSTMENT -- Oxamyl, 3-hydroxycarbo-
                furan, and carbaryl can all degrade quickly in water held at
                room temperature.6'7  This short term degradation is of
                concern during the time samples are being shipped and the
                time processed samples are held at room temperature in
                autosampler trays.  Samples targeted for the analysis of
                these three analytes must be preserved at pH 3.   The pH
                adjustment also minimizes analyte biodegradation.
                Preservation study results given in Table 10 indicate that
                method analytes are stable in water samples for at least 28
                days when stored using the following conditions:

                8.2.1     The  samples  must be  iced  or refrigerated at  4°C  from
                          the  time of  collection  until  extraction.

                8.2.2     Add  0.6 mL of  monochloroacetic  acid  buffer  to  the
                          60-mL sample bottle.  Add buffer to  the  sample
                          bottle at the  sampling  site or  in the laboratory
                          before shipping to the  sampling site.

                8.2.3     After sample is collected in  bottle  containing
                          buffer,  seal the sample bottle  and shake vigorously
                          for  1 min.
9.   CALIBRATION
         9.1    Establish HPLC operating conditions equivalent to those
                indicated in Table 2.  Calibrate the HPLC system using the
                internal standard technique (Section 9.2).

         9.2    INTERNAL STANDARD CALIBRATION PROCEDURE.   The analyst must
                select one or more internal standards similar in analytical
                behavior to the analytes of interest.  The  analyst must
                further demonstrate that the measurement of the internal
                standard is not affected by method or matrix interferences.
                BDMC has been identified as a suitable internal standard.

-------
                9.2.1    Prepare calibration standards at a minimum  of  three
                         (suggested five) concentration levels  for each
                         analyte of interest by adding volumes  of one or more
                         stock standards to a volumetric flask.  To  each
                         calibration standard, add a known constant  amount of
                         one or more internal standards, and dilute  to  volume
                         with buffered reagent water.  To prepare buffered
                         reagent water, add 10 mL of 1.0 M monochloroacetic
                         acid buffer to 1 L of reagent water.   One of the
                         standards should be representative of  an analyte
                         concentration near, but above, the MDL.  The other
                         concentrations should correspond to the range  of
                         concentrations expected in the sample  concentrates,
                         or should define the working range of  the detector.

                9.2.2    Inject 400 uL of each calibration standard  and
                         tabulate the relative response for each analyte
                         (RRa)  to the internal standard using the equation:

                         RRa =  Aa/Als

                         where:    Aa   =  the peak area of the  analyte,  and
                                   Als =  the peak area of the  internal
                                          standard.

                         Generate a calibration curve of analyte RRa versus
                         analyte concentration in the sample in ug/L.

                9.2.3    The working calibration curve must be  verified on
                         each working shift by the measurement  of one or more
                         calibration standards.  If the response for any
                         analyte varies from the predicted response  by  more
                         than 10%, the test must be repeated using a fresh
                         calibration standard.  Alternatively,  a new
                         calibration curve must be prepared for that analyte.
10.       QUALITY CONTROL
         10.1   Each laboratory using this method is required to operate a
                formal quality control (QC) program. The minimum requirements
                of this program consist of the following:  an initial
                demonstration of laboratory capability;  the analysis of QC,
                laboratory control,  and spiked samples as a continuing check
                on recovery performance;  the analysis of field duplicates as
                a continuing check on precision;  the analysis of reagent
                blanks as a continuing check on contamination; and frequent
                analysis of the instrument QC standard to assure acceptable
                instrument performance.

         10.2   INITIAL DEMONSTRATION OF CAPABILITY -- To establish the
                ability to perform this method, the analyst must perform the
                following operations.

-------
       10.2.1    Select a representative spike concentration for each
                 analyte to be measured.   Using stock standards that
                 differ from calibration standards,  prepare a
                 laboratory control check sample concentrate in
                 methanol 1000 times more concentrated than the
                 selected spike concentrations.

       10.2.2    Using a syringe,  add 50 uL of the check sample
                 concentrate to each of a minimum of four 50-mL
                 aliquots of reagent water.   Add 0.5 mL of 1.0 M
                 monochloroacetic  acid buffer to each spiked sample.
                 A representative  ground water may be used in place
                 of the reagent water,  but one or more unspiked
                 aliquots must be  analyzed to determine background
                 levels, and the spike level must exceed twice the
                 background level  for the test to be valid.   Analyze
                 the aliquots according to the method beginning in
                 Section 11.

       10.2.3    Calculate the average percent recovery (R),  and the
                 standard deviation of the percent recovery (s),  for
                 the results.  Ground water background corrections
                 must be made before R and s calculations are
                 performed.

       10.2.4    Tables 3-9 provide single operator recovery and
                 precision data obtained for the method analytes from
                 reagent and artificial ground waters.   Similar
                 results should be expected from reagent water for
                 all compounds listed in the method.   Compare the
                 results obtained  in section 10.2.3 to the values
                 listed in Tables  3-9.   If the results are not
                 comparable, review potential problem areas  and
                 repeat the test.

10.3   In recognition of the rapid advances occurring in chromato-
       graphy,  the analyst  is permitted certain options to improve
       the separations  or  lower the cost of measurements.  Each time
       such modifications  to the method are made, the analyst is
       required to repeat  the procedure in Section 10.2.

10.4   ASSESSING ANALYTE RECOVERY

       10.4.1    Each quarter,  it  is essential that the laboratory
                 analyze (if available) QC check standards for each
                 contaminant.  If  the criteria established by USEPA
                 and provided with the QA standards are not  met,
                 corrective action needs  to be taken and documented.

       10.4.2    After every 10 samples,  and preferably in the middle
                 of each day, analyze a laboratory control standard.
                 Calibration standards may not be used for accuracy
                 assessments, and  the laboratory control standard may

-------
         not be used  for calibration of  the  analytical
         system.

         10.4.2.1  Laboratory control standard concentrate --
                   Internally prepared laboratory control
                   standards are used to provide the routine
                   assessment of recovery.  They should be
                   prepared from a separate set of stock
                   standards, i.e., different from calibra-
                   tion standards.

         10.4.2.2  Laboratory control standard -- Add an
                   appropriate aliquot of the control
                   standard concentrate to 50 mL of buffered
                   reagent water in a 50 mL volumetric flask.

         10.4.2.3  Analyze the 50 mL laboratory control
                   standard as described in Section 11.   For
                   each analyte in the laboratory control
                   standard, calculate the percent recovery
                   (Pi)  using the equation:

                           100S1
                      PI -  ------
                     where:   Si =  the  analytical  result  from
                                   the  laboratory  control
                                   standard in ug/L,  and

                               Tx =     the known spiked
                                       concentration of the
                                       laboratory control
                                       standard in ug/L.

10.4.3    It  is  essential  that the  laboratory  analyze an
          unknown performance  evaluation  sample (when
          available)  once  per  year  for  all  regulated
          contaminants measured.  Results need to be within
          acceptable  limits  established by  USEPA for each
          analyte .

10.4.4    The  laboratory  is  required  to collect in duplicate  a
          portion of  the  samples  to monitor spike  recoveries.
          The  frequency of spiked sample  analysis  must  be at
          least  10  percent of  all samples,  or  one  spiked
          sample per  month,  whichever is  greater.  One  aliquot
          of  the sample must be spiked  and  analyzed as
          described in Section 11.

-------
10.5   ASSESSING ANALYTE PRECISION

       10.5.1    Precision assessments for this method are based upon
                 the analysis  of field duplicates.   Analyze field
                 duplicates for at least 10 percent of all samples.
                 To the  extent practical,  the duplicated sample
                 should  contain reportable levels of method analytes.

       10.5.2    For each analyte in each duplicate pair,  calculate
                 the relative  range (RR1)  using the equation:

                       100R,
                 RR,  =	
                         X,

                 where:    Rj =     the absolute difference between
                                   the duplicate measurements Xj^  and
                                   X2  in ug/L,  and

                          Xx =     the average concentration found
                                   ([X:  +  X2]/2)  in ug/L.

       10.5.3    Individual relative range measurements are pooled to
                 determine the average relative range or to develop
                 an expression of relative range as a function of
                 concentration.

10.6   ASSESSING CONTAMINATION

       10.6.1    Laboratory reagent blanks -- Before processing any
                 samples,  the  analyst must demonstrate that all
                 glassware and reagent interferences are under
                 control.   This is accomplished by  the analysis of a
                 laboratory reagent blank.  A laboratory reagent
                 blank  is a 50 mL aliquot of reagent water analyzed
                 as if  it were a sample.  Each time a set of samples
                 is extracted  or there is a change  in reagents, a
                 laboratory reagent blank must be processed to assess
                 laboratory contamination.  If the  reagent blank
                 exhibits a peak within the retention time window of
                 any analyte which is greater than  or equal to one-
                 half the MDL  for that analyte, determine source  of
                 contamination before proceeding.

10.7   ASSESSING INSTRUMENT PERFORMANCE -- The  laboratory  is
       required to demonstrate the ability to  generate acceptable
       HPLC performance for each analyte.  Analyte peak shapes
       comparable to those shown in Figure 1 should be
       demonstrated.  Instrument performance should be monitored on
       a daily basis by analysis of the instrument QC  standard.

-------
                10.7.1    Instrument QC standard --  The instrument QC
                          standard contains compounds designed to indicate
                          appropriate instrument sensitivity,  column
                          performance,  and chromatographic performance.
                          Instrument QC standard components and performance
                          criteria are  listed in Table 11.  Inability to
                          demonstrate acceptable instrument performance
                          indicates the need for reevaluation of the HPLC-UV
                          system.   An HPLC-UV chromatogram generated from the
                          analysis of the instrument QC standard is shown in
                          Figure 3.

                ANALYTE CONFIRMATION -- When doubt exists over the  identi-
                fication of a peak on the chromatogram, confirmatory
                techniques such as chromatography with a  dissimilar column,
                or ratio of absorbance at two or more wavelengths must be
                used.  A suggested confirmation column is described in Table
                2.

         10.9   ADDITIONAL QC --  It  is recommended that the laboratory adopt
                additional quality assurance practices for use with this
                method.  The specific practices that are  most productive
                depend upon the needs of the laboratory and the nature of the
                samples.

11-       PROCEDURE

         11.1   PH ADJUSTMENT AND  FILTRATION

                11.1.1    Adjust the pH of the  sample or standard to pH  3 +
                          0.2  by adding 0.5 mL of 1.0 M monochloroacetic acid
                          buffer to each 50 mL of sample.   This step is  not
                          necessary if  sample pH was adjusted during sample
                          collection as a preservation precaution.   Fill a
                          50-mL volumetric flask to the mark with the sample.
                          Add  5 uL of the internal standard spiking solution
                          and  mix by inverting the flask several times.

                11.1.2    Affix the three-way valve to a 10-mL syringe.   Place
                          a clean filter in the filter holder and affix  the
                          filter holder and the 7- to 10-cm syringe needle to
                          the  syringe valve.  Rinse the needle and syringe
                          with reagent  water.  Prewet the filter by passing 5
                          mL of reagent water through the filter.  Empty the
                          syringe and check for leaks.   Draw 10 mL of sample
                          into the syringe and expel through the filter.  Draw
                          another 10 mL of sample into the syringe, expel
                          through the filter, and collect the last 5 ml  for
                          analysis.  Rinse the syringe with reagent water.
                          Discard the filter.

-------
         11.2   LIQUID CHROMATOGRAPHY

                11.2.1    Table  2  summarizes  the recommended operating
                          conditions  for  the  liquid chromatograph.   Included
                          in Table 2  are  retention times observed using this
                          method.   An example of the separations achieved
                          using  these conditions is shown in Figure 1.  Other
                          HPLC columns, chromatographic conditions, or
                          detectors may be  used if the requirements of Section
                          10.2 are met.

                11.2.2    Calibrate the system daily as described in Section
                          9.  The  standards and extracts must be in pH3
                          buffered water.

                11.2.3    Inject 400  uL of  the sample extract.   Record the
                          volume injected and the resulting peak size in area
                          units.

                11.2.4    The width of the  retention time window used to make
                          identifications should be based upon measurements  of
                          actual retention  time variations of standards over
                          the course  of a day.   Three times the standard
                          deviation of a  retention time can be used to
                          calculate a suggested window size for a compound.
                          However,  the experience of the analyst should weigh
                          heavily  in  the  interpretation of chromatograms.

                11.2.5    If the response for the peak exceeds the  working
                          range  of the system,  dilute the extract and
                          reanalyze.

12.       CALCULATIONS

         12.1   Calculate analyte  concentrations in the sample from the
                relative response  for the analyte  (RRa) to the internal
                standard using the equation the calibration curve described
                in Section 9.3.2.

         12.2   For samples processed as  part of a set where the laboratory
                control standard recovery falls outside of the control limits
                in Section 10.4, data for the affected analytes must be
                labeled as suspect.


13.       PRECISION AND ACC1 »_CY

         13.1   In a single laboratory,  analyte recoveries from reagent water
                were determined at five  concentration levels.  Results were
                used to determine  analyte EDLs and demonstrate method range.
                EDL determination  results are given in Table 3.  Method range
                results are given  in Tables 4-7.

-------
         13.2   In a single laboratory, analyte recoveries from two
                artificial ground waters were determined at one concentration
                level.  Results were used to demonstrate applicability of the
                method to different ground water matrices.  Analyte
                recoveries from the two artificial matrices are given in
                Tables 8 and 9.

         13.3   In a single laboratory, analyte recoveries from a ground
                water preserved by adjusting to pH 3 with monochloroacetic
                acid buffer were determined 0, 14, and 28 days after sample
                preparation.  Samples were stored at 4°C or -10°C and were
                protected from light.  Results were used to predict expected
                analyte stability in ground water samples.  Analyte
                recoveries from the preserved, spiked ground water samples
                are given in Table 10.

REFERENCES

1.      Moye,  H.A.,  S.J.  Sherrer,  and P.A.  St.  John,  "Dynamic  Labeling of
        Pesticides for High Performance Liquid Chromatography:  Detection of
        N-Methylcarbamates  and o-Phthalaldehyde,"  Anal.  Lett.  10,  1049,  1977.

2.      ASTM Annual  Book of Standards,  Part 31,  D3694,  "Standard  Practice for
        Preparation  of Sample  Containers and for Preservation," American
        Society for  Testing and Materials,  Philadelphia,  PA, p. 679,  1980.

3.      "Carcinogens -  Working with Carcinogens,"  Department of Health,
        Education, and Welfare,  Public Health Service,  Center  for  Disease
        Control,  National Institute for Occupational  Safety and Health,
        Publication  No.  77-206,  Aug.  1977.

4.      "OSHA Safety and Health Standards,  General Industry,"  (29  CFR 1910),
        Occupational Safety and Health Administration,  OSHA 2206,  (Revised,
        January 1976) .

5.      "Safety in Academic Chemistry Laboratories,"  American  Chemical Society
        Publication,  Committee on Chemical  Safety,  3rd Edition, 1979.

6.      Foerst, D.C.  and H.A.  Moye,  "Aldicarb in Drinking Water via Direct
        Aqueous Injection HPLC with Post Column Derivatization,"  Proceedings
        of the 12th  annual  AWWA Water Quality Technology conference,  in press
        1985.

7.      Hill,  K.M.,  R.H.  Hollowell,  and L.A.  DalCortevo,  "Determination of
        N-Methylcarbamate Pesticides in Well Water by Liquid Chromatography
        and Post Column Fluorescence Derivatization," Anal. Chem,  56,  2465
        (1984) .

8.      ASTM Annual  Book of Standards, Part 31, D3370,  "Standard  Practice for
        Sampling Water," American Society for Testing and Materials,  Philadel-
        phia,  PA,  p.76,  1980.

-------
             TABLE 1.  METHOD ANALYTES


Analyte
Aldicarb
Aldicarb sulfone
Aldicarb sulfoxide
Baygon
Carbaryl
Carbofuran
3-Hydroxycarbofuran
Methiocarb
Me thorny 1
Oxamyl
Chemical
Abstracts Service
Registry Number
116-06-3
1646-88-4
1646-87-3
114-26-1
63-25-2
1563-66-2
16655-82-6
2032-65-7
16752-77-5
23135-22-0

Ident .
Code(a)
6
2
1
8
9
7
5
10
4
3
(a)   Code used for identification of peaks in figures;
     IS = internal standard.

-------
 TABLE  2.   PRIMARY AND  CONFIRMATION  CHROMATOGRAPHIC  CONDITIONS
                             Retention Time for Given Conditions
Analyte
     Primary (a)(c)
Confirmation (b)(d)
Aldicarb
Aldicarb sulfone
Aldicarb sulfoxide
Baygon
Carbaryl
Carbofuran
3 -Hydroxycarbof uran
Methiocarb
Methomyl
Oxamyl

0.429
0.421
0.834
0.867
0.824
0.657
0.984
0.518
0.489
0.761 21.4
12.2
17.5
23.4
25.4
24.4
19.0
28.6
14.8
14.6
(a)  Retention time relative to BDMC internal standard which elutes
     at 35.5 min.
(b)  Absolute retention time in minutes.
(c)  Primary conditions:

             Column:  250 mm x 4.6 mm I.D. Altex Ultrasphere ODS
                      (5 urn)
       Mobile phase:  Linear gradient from 15:85 methanol:water to
                      methanol in 32 min
          Flow rate:  1.0 mL/min
   Injection volume:  400 uL
           Detector:  Fluorescence; excitation 230 nm; emission
                      418 nm

(d)  Confirmation conditions:
             Column:
       Mobile phase:

          Flow rate:
   Injection volume:
           Detector
250 mm x 4.6 mm I.D. Supelco LC-1 (5 urn)
Linear gradient from 15:85 methanol:water to
methanol in 32 min
1.0 mL/min
400 uL
Fluorescence; excitation 230 nm; emission
-+18 nm

-------
TABLE 3.   RECOVERIES OF ANALYTES FROM REAGENT WATER (SPIKING LEVEL 1) AND EDLs (a)
Spiking Amt in
Level, Blank,


Analyte ug/L ug/L n(b) R(c)
Aldicarb 1.0 ND (g)
Aldicarb sulfone 2.0 ND
Aldicarb sulf oxide 2.0 ND
Baygon 1.0 ND
Carbaryl 2.0 ND
Carbofuran 1.5 ND
3-Hydroxycarbofuran 2.0 ND
Methiocarb 4.0 ND
Methomyl 0.50 ND
Oxamyl 2.0 ND
(a) Amounts corrected for levels detected in
(b) n = number of recovery data points.
(c) R = average percent recovery.
(d) S = standard deviation.
(e) RSD = percent relative standard deviation
8
8
8
7
8
7
8
8
7
8
blank




(f) EDL = estimated detection limit in sample
107
83
47
101
97
90
108
82
102
82





in ug/L

S(d) RSD(
0.0728
0.337
0.196
0.323
0.443
0.166
0.626
0.638
0.0931
0.287





; calculated


e) EDL(f)
7
20
21
32
23
12
29
19
18
17





by
1.0
2.0
2.0
1.0
2.0
1.5
2.0
4.0
0.50
2.0






          multiplying standard deviation (S)  times  the  students'  t value  appropriate
          for  a 99%  confidence level  and a  standard deviation estimate with n-1
          degrees  of freedom,  or a level of compound in sample yielding a peak in
          the  final  extract with signal-to-noise  ratio  of  approximately 5, whichever
          value is higher.
(g)  ND = interference not detected in blank.

-------
TABLE 4.   RECOVERIES OF ANALYTES FROM REAGENT WATER (SPIKING LEVEL 2) (a)
Spiking Amt in
Level, Blank,
Analyte
Aldicarb
Aldicarb
Aldicarb
Baygon
Carbaryl


sulfone
sulfoxide


Carbofuran
3-Hydroxycarbofuran
Methiocarb
Me thorny 1
Oxamyl


(a) Amounts corrected
(b) n -
(c) R =
(d) S =
(e) RSD
(f) ND
ug
2
4
4
2
4
3
4
8
1
4
for
number of recovery
/L
.0
.0
.0
.0
.0
.0
.0
.0
.0
.0
amount
data
ug/L
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
found
points
n(b)
(f) 8
8
8
8
8
8
8
7
8
8
in blank.

R(c)
113
100
73
97
94
93
93
80
76
88



0
0
0
0
0
0
0
0
0
0


S(d)
.125
.251
.283
.181
.292
.151
.392
.246
.0893
.246


RSD(e)
6
6
10
9
8
5
11
4
12
7


average percent recovery.
standard deviation.
= percent relative
= interference
not
standard deviation.
detected in
blank.





-------
TABLE 5.  RECOVERIES OF ANALYZES FROM REAGENT WATER (SPIKING LEVEL 3) (a)
Spiking Amt in
Level, Blank,

Analyte ug/L ug/L n(b)
Aldicarb 5.0 ND (f)
Aldicarb sulfone 10 ND
Aldicarb sulfoxide 10 ND
Baygon 5.0 ND
Carbaryl 10 ND
Carbofuran 7.5 ND
3-Hydroxycarbofuran 10 ND
Methiocarb 20 ND
Me thorny 1 2.5 ND
Oxamyl 10 ND
8
8
8
8
8
8
8
7
8
8

R(c)
115
101
97
106
97
102
102
94
105
100


S(d) RSD(e)
0.172
0.407
0.441
0.152
0.607
0.346
0.386
0.453
0.0951
0.423
3
4
5
3
6
5
4
2
4
4
(a) Amounts corrected for amount found in blank.
(b) n = number of recovery data points
(c) R = average percent recovery.
(d) S = standard deviation.












(e) RSD = percent relative standard deviation.
(f) ND = interference not detected in blank.





-------
TABLE 6.  RECOVERIES OF ANALYTES FROM REAGENT WATER (SPIKING LEVEL 4)  (a)
Spiking Amt in
Level, Blank,
Analyte ug/L ug/L
Aldicarb 10 ND (f)
Aldicarb sulfone 20 ND
Aldicarb sulfoxide 20 ND
Baygon 10 ND
Carbaryl 15 ND
Carbofuran 15 ND
3-Hydroxycarbofuran 20 ND
Methiocarb 40 ND
Me thorny 1 5.0 ND
Oxamyl 20 ND
(a) Amounts corrected for amount found in
(b) n = number of recovery data points
(c) R = average percent recovery.
(d) S = standard deviation.


n(b)
8
8
8
8
8
8
8
8
8
8
b 1 ank .





R(c)
105
91
92
94
112
96
92
83
96
90






S(d)
0.300
0.657
0.441
0.309
0.298
0.247
0.910
0.722
0.0912
0.501






RSD(e)
3
4
2
3
2
2
5
2
2
3




(e) RSD = percent relative standard deviation.
(f) ND = interference not detected in blank.

-------
TABLE 7.  RECOVERIES OF ANALYTES FROM REAGENT WATER (SPIKING LEVEL 5) (a)
Spiking Amt in
Level ,
Analyte
Aldicarb
Aldicarb sulfone
Aldicarb sulfoxide
Baygon
Carbaryl
Carbofuran
3 -Hydroxycarbofuran
Methiocarb
Me thorny 1
Oxamyl
(a) Amounts corrected
Blank,
ug/L ug/L
25
50
50
25
50
38
50
100
13
50
for
(b) n = number of recovery
ND (f)
ND
ND
ND
ND
ND
ND
ND
ND
ND
amount found in
data points

n(b)
8
8
8
8
8
8
8
8
8
8
blank.


R(c)
98
92
96
91
83
91
90
82
98
89




1
2
2
1
1
1
2
5
0
1



S(d)
.31
.06
.76
.03
.76
.82
.00
.85
.529
.80



RSD(e)
5
4
6
4
4
5
4
7
4
4


(c) R = average percent recovery.
(d) S = standard deviation.
(e) RSD = percent relative
(f) ND = interference
not
standard deviation.
detected in blank.

-------
TABLE 8.  RECOVERIES OF ANALYTES FROM HARD ARTIFICIAL GROUND WATER
          (SPIKING LEVEL 3) (a)
Spiking Amt in
Level, Blank,

Analyte ug/L ug/L n(b)
Aldicarb 5.0 ND (f)
Aldicarb sulfone 10 ND
Aldicarb sulfoxide 10 ND
Baygon 5.0 ND
Carbaryl 10 ND
Carbofuran 7.5 ND
3-Hydroxycarbofuran 10 ND
Methiocarb 20 ND
Methomyl 2.5 ND
Oxamyl 10 ND
8
8
8
8
8
8
8
8
8
8

R(c)
106
98
105
96
94
102
98
102
98
97


S(d) RSD(e)
0.177
0.441
0.393
0.224
0.454
0.245
0.494
0.856
0.0863
0.269
3
4
4
5
5
3
5
4
4
3
(a) Amounts corrected for amount found in blank.
(b) n = number of recovery data points
(c) R = average percent recovery.
(d) S = standard deviation.
(e) RSD = percent relative standard deviation
(f) ND = interference not detected in blank.





















-------
TABLE 9.  RECOVERIES OF ANALYTES FROM ORGANIC-CONTAMINATED ARTIFICIAL
          GROUND WATER (SPIKING LEVEL 3) (a)

Analyte
Aldicarb
Aldicarb
Aldicarb
Baygon
Carbaryl

Spiking
Level ,
Amt in
Blank,

ug/L ug/L n(b)

sulfone
sulfoxide


Carbofuran
3 -Hydroxycarbof uran
Methiocarb
Me thorny 1
Oxamyl


5
10
10
5
10
7.5
10
20
2.5
10
(a) Amounts corrected for
(b) n -
(c) R =
(d) S =
(e) RSD
(f) ND
number of
recovery
ND (f)
ND
ND
ND
ND
ND
ND
ND
ND
ND
8
8
8
7
8
7
8
7
8
8

R(c)
102
95
94
97
104
100
101
112
105
102



S(d) RSD(e)
0
0
1
0
1
0
0
0.
0.
1.
.406
.981
.05
.300
.08
.524
.969
.660
.244
.03
8
10
11
6
10
7
10
3
9
10
amount found in blank.
data points





average percent recovery.
standard deviation.
= percent
relative
= interference not
standard deviation
detected in blank.











-------
                TABLE  10.   RESULTS FROM METHOD  5 PRESERVATION STUDIES  (c)
Spiking
Day 0
Day 14 (
:-10°C)
Day 14
(4°C)
Day 28
(-10°C) Day 28 (4°C)
Level,
Analyte
Aldicarb
Aldicarb sulfone
Aldicarb sulfoxide
Baygon
Carbaryl
Carbofuran
3-Hydroxycarbofuran
Methiocarb
Methomyl
Oxamyl
g/L
5.0
10
10
5.0
10
7.5
10
20
2.5
10
R(a)
100
99
100
98
100
100
95
110
110
98
RSD(b)
9
9
9
10
6
9
9
4
12
6
R
100
93
91
91
92
95
89
100
90
85
RSD
4
3
6
2
4
3
6
1
6
4
R
110
99
100
100
95
110
100
98
96
95
RSD
2
3
2
4
3
7
3
1
5
5
R
100
97
98
88
99
95
100
99
93
100
RSD
6
0
2
2
18
2
8
0
2
11
R
83
86
91
93
89
93
95
94
96
94
RSD
1
8
5
9
5
3
11
6
5
9
(a)  R = average percent recovery.
(b)  RSD = percent  relative standard deviation.
(c)  Storage temperature given in parentheses.

-------
                                 TABLE  11.    QUALITY CONTROL STANDARD
Test
Sensitivity
Chromatographic performance
Column performance
Analyte
3- Hydroxycarbof uran
Aldicarb sulfoxide
Methiocarb
4-Bromo-3,5-dimethylphenyl
N-methylcarbamate (IS)
Cone,
9/L
2
100
20
10
Requi rements
Detection of analyte; S/N > 3
0.95 1 .0 (c)
(a)  PSF  =  peak symmetry factor.   Calculated  using the equation:
               w(1/2>
       PSF  =  ...........
             0.5 x W(1/2)
     where  w(1/2) is the width of the front of the peak at half height  and UC1/2) is the
     peak width at half height.
(b)  PGF  =  peak Gaussian factor.   Calculated  using the equation:
             1.83 x W(1/2)
       PGF  =  - ....... -----
     where W(1/2) is the peak width at  half height and U(1/10)  is  the  peak width at tenth
     height.
(c)   Resolution between the two peaks as  defined by the equation:
            t
       R  = ---
            U
     where t is the difference in elution times between the two peaks  and W is the average
     peak width, at the baseline,  of the  two peaks.

-------
                                Appendix B
                                Revision No. 2
                                Date:  July 29, 1990
                                Page 1 of 5
              APPENDIX B

NPS ANALYTE REPORTING BELOW MRL AND
      IDENTIFYING UNKNOWN PEAKS

-------
          |    UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
                                CHCMwmr L*«MIATO*Y. NASA/NCTL
                           •UILMMG UN, NTO. M«M*Sim «0
                               June 1, 1988
MEMORANDUM

SUBJECT:  NFS Analyte Reporting Below MRL and Identifying Unknown Peaks

FROM:     Bob Maxey/ Analytical Coordinator
          Environmental Chemistry Laboratory

TO:       Dave Munch, Analytical Coordinator
          TSD-Cincinnati

          Atfcry E. Dupuy, Jr., Technical Monitor
          Environmental Chemistry Laboratory

     Attached are the procedures that NPS analytical contractors and referee
laboratories must adhere to in complying with the OPP request to report the
presence of analytes below the Minimal Reporting Limits and to attempt identi-
fication of unknown peaks.  Please see that your contractors and Technical
Monitors get this information and that applicable parts are incorporated into
their respective QAPPs.

     If you have any questions, give me a call.

-------
        Determining and Reporting the Presence of NPS Analytes Below The
             Minimal Reporting Levels and Identifying Unknown Peaks


Background Information

     The Office of Pesticide Programs (OPP) has requested that the NFS analyti-
cal contractors and referee laboratories make an effort to report the presence
of NPS analytes below the Minimal Reporting Levels (MRL).  We have also been re-
quested to attempt to identify unknown peaks or responses.  To assure that spur-
ious or ambiguous data is not reported and that a uniform system or analytical
routine is used at all laboratories to accomplish these requests/ criteria have
been developed for handling both situations.

Procedure for Determining and Reporting the Presence of NPS Analytes Below the
MRL

1.   For methods 1-7, only peaks with responses of between one-half the establis
     ed MRL and the MRL V on the primary column will be investigated.  A respon
     on the "secondary" GC column, indicating the presence of the analyte, is al
     required for additional work.

2.a  The first occurrence of a peak meeting the requirements of  (1) is noted and
     reported to the Technical Monitor, but no action is taken V.  Upon a secon
     occurrence of the same suspect analyte/ additional work is  required as fol-
     lows.  After five successive failures to "confirm" on the secondary column
     the response on the primary column/ discussions with OPP personnel will tafc
     place before continuing low-level analytical work on the analyte(s).

  b  With methods 1,2,3, 6 and 7, for responses meeting the requirements of (1) .
     and  (2), the laboratory will attempt LR GC/MS c/ confirmation if the GC/MS
     analyst feels it is within the capability of his instrument.  If the conf ir
     mat ion is not within the capability of the laboratory/ such extracts are
     sent weekly, under iced conditions by next-day air/ to the  appropriate re-
     feree  laboratory having HR GC/MS c/ capabilities.  Copies of chromatograms
     and all pertinent sample information must be sent along with the extracts
     including extracts of the related Method Blank.   (NPS will  absorb the cost
     of these shipments.)  It is preferred that extracts be in sealed glass
     ampules/ but other vials and teflon-faced closures are acceptable if  they
     provide a tight seal and do not contribute interferences to the extracts.
     Volume  level must be marked on the outside of the vial or ampule.


A/ » NPS method 1 MRL « 4 x EDL     NPS method 5 MRL  - 3  x H>L
     NPS method 2 MRL « 5 x H)L     NPS method 6 MRL  = 3  x EDL
     NPS method 3 MRL • 5 x HDL     NPS method 7 MRL  - 3  x H)L
     NPS method 4 MRL - 3 x EDL

B/ « Method 6  has an MRL > the Health Advisory Level. All suspect  ETU
     responses of 1/2 MRL - MRL  require additional work  for  this method.

C/ - LR - GC/MS - Low Resolution mass spectronetry.
     HR GC/MS  » High Resolution  mass spectrometry.

-------
  c.  For Methods 4 and 5, HPLC Methods/  there  is  no provision for GC/MS
      confirmation.  Suspect analytes between 1/2  MRL - MRL will be subject
      to (1) and (2a) above.

      Provisions of (2b) also apply except references to GC/MS requirements.

3.    Whether the identification of the analyte is attempted at the contractor
      laboratory or at the referee laboratory,  only analytes positively con-
      firmed by GC/MS will be reported beyond the  Technical Monitor for the
      Method and the Analytical Coordinators.   No  unconfirmed data will be
      reported outside the NFS analytical system.   Unsuccessful attempts  at
      confirmation will also be reported  to the Technical Monitor.

4.    Following either the successful GC/MS confirmation of two such responses
      for the same analyte or two successive failures to confirm the analyte
      by GC/MS without any prior successful GC/MS  confirmation on any samples,
      discussions with OPP personnel will take  olace before continuing low
      level analytical work on that analyte.

Procedure for Determining the Identity of Son-SPS  Analytes

      It is expected that, over the course of  the  NPS Program, numerous
extraneous  responses will be evident on  chromatograms from the various methods.
The contractor or referee laboratories will be  required to attempt identifi-
cation of peaks or responses on the primary column exhibiting the minimal criter
below.

      1.  For Methods 1, 2, 3, 6, and 7,  if, upon  initial analyses, the response
          of an extraneous peak on the primary  column is equal to or greater  thai
          the response of the nearest NPS analyte on that column at 10 x  MRL
          (Minimal Reporting Level), an attempt must be made to identify  that un-
          known peak or response by GC/MS.  Full scan spectra and subsequent
          library search are expected and must  be followed by comparison  of  the
          spectra of the unknown compound with  those of an authentic standard of
          the suspected compound.

      2.  The work in (1) must be attempted by  the contractor and/or referee  lab
          oratories on the first occurrence of such a peak and the results of th
          attempt reported to the Technical Monitor for the Method.  If the
          analytical contractor feels  his system or instrument is not capable of
          the confirmatory work, he must  send both that extract and that  of the
          related Method Blank to the  appropriate referee lab under iced
          conditions by next-day air.

          It is preferred that extracts  be in sealed glass ampules, but other
          vials and teflon-faced closures are acceptable if they provide a tight
          seal and do not contribute interferences to the extracts.  Volume lev*
          must be marked on the outside  of the vial or ampule.   (NPS will absorb
          coats of these shipments.)

-------
      Specific aarople and analytical information  must accompany each auch
      extract.

        o  Sample i.d. number,  weight of sample matrix contained in the
           ampule, copies of chrotnatograms from the primary GC coluan/
           identification of the retention window for the unknown response(s)
           as defined by the last NPS analyte to  elute before the unknown
           peak or response and the first NPS analyte to elute following the
           unknown response.

3.    Whether the identification of the unknown compound is attempted at the
      Contractor Laboratory or at the referee laboratory/ only the compounds
      positively confirmed by GC/MS will be reported beyond the Technical
      Monitor for the Method and the Analytical Coordinators.  No unconfirmed
      data will be reported outside the NPS analytical system.  Unsuccessful
      attempts at identification will also be reported to the Technical
      Monitor.

4.    Following either the successful confirmation of two such extraneous peaks
      proving to be the same compound or two failures to identify a response
      with the same retention time without a prior successful QC/MS confirmation
      on a sample, discussions with OPP personnel will take place before continua
      with identification work on that particular compound.

      THE QUALITY ASSURANCE PROJECT PLANS FOR BOTH  THE ANALYTICAL CONTRACTORS
      AND REFEREE LABORATORIES FOR METHODS 1, 2,  3, 6, AND 7 MUST REFLECT THEIR
      COMMITMENTS TO THESE TWO REQUIREMENTS.

      THE QUALITY ASSURANCE PROJECT PLANS FOR BOTH THE ANALYTICAL CONTRACTORS
      AND REFEREE LABORATORIES FOR METHODS 4 AND  5 MUST REFLECT THEIR COMMIT-
      MENTS TO THE REQUIREMENT FOR DETERMINING AND REPORTING NFS ANALYTES BELOW
      THE MRL.

-------
                                 Appendix C
                                 Revision No. 2
                                 Date: July 29, 1990
                                 Page 1 of 3
               APPENDIX C

CHANGES IN NPS LABORATORY PROCEDURES

-------
if-—-*
        I    UNITED STATES ENVIRONMENTAL PROTECTION AGENCY
                                CINCINNATI OHIO  45268
   KEXORANDUX

   DATE:    July 14. 1988

   SUBJECT: Changes in NPS Laboratory Procedures

   TKOX:    David J. Hunch. TS2 Project Manager
            National Pesticide Survey

   TO:      R?S Technical Heritors  (See below;
       T>.c- following rir.cr changes in laboratory operations  are  beir.g -« = >..

       1.   Spiking Levels 'Methods I-"'

            Currently. selected !.*"3 sartles are being spined  at  either Lev«l
       :  (5 ti.-ts M?L  . Lev;l : ;i: ti-bs MF.L  . cr Level  3  (2:  tires  M-l .
       In rany cases,  spiking at Level  2 has crested analyt-:  concer.traticn;
       in sarples whirr, exceed tr.r linear range cf the instrumentation.   Any
       Level 3 spiked  sample's currer.tly or. har.i should be analyred:  howeve:,!
       no further requests v-.;i be cad; to spir.e sarples  at  Level  ?.

            In order tc .raintair. three spihing levels, a  Level  0 *2  tir.ss
       XRL) is bein; adcei.  Ls.::tat:ry Cor.trsl Standards and Tire  Storace
       Sa-;les,are to  contin-c tc be ssiked at Level 2  (1C  tires MHL' .

       2.   Spiking Levels (Method 9)

            Currently, sairple spiking levels used for Method  9  are,  Level '.
        (2 times MRL) .  Level 2 (10 times HRL) ,  and Level 3 (10.000  ug/L) .  Tl..
       spiking levels  art to reaain the sars;  however. Level  0  will  now  it  I
       tir.es MRL, Level 1 10 tices HRL, and Level 3 10,000  ug/L.


       3.   Data Reporting Forrat

            In order for the data reporting forcat to satch  the  rec.uirerer.ts
       for reporting suspected NFS analytes observed on the  prirary  colurr. ,
       at a concentration between 1/2 HRL and  HRL  (see ceaorandun  entitled
       "Determining and Reporting the Presence of NTS Analytes  below  thi
       Hiniaui Reporting Levels and Identifying Unknown Peaks,"  by Bob Haxey
       fi/1/88), further clarification is required.  In those  cases  where tl.c
       presence of an  NFS analyte at a  concentration between  1/2 HRL  and t.h~
       HRL is successfully confirmed, the primary and conf irrcational  colur:.
       data for that analyte should be  reported as "-111".   In  those  cases
       where conf irmational analyses are either not required, or the
       conf irmational  analyses did not  confirm the presence  of  the  analyte,
       the primary column data for that analyte should be reported  as "-222".

-------
                                    -2-
    Plcase transmit this information  to bot-  your  contract  and refere;
laboratories, as soon as possible.  If you  have  any questions concern;
these ir«»s, please let T.e know.
Addressees:

    A. Dupuy
    L. Kacphake  (TSD
    C. "adding  (TSr;
    R. Maxey  (OP?)
    K. Sorrel!  'TST1
    R. Tho-as  (TS:1
    H. Brass  (TS:
    C. Freebis  (CSC
    A. Kroner  (?s:

-------
                                       Appendix D
                                       Revision No. 2
                                       Dale: July 29, 1990
                                       Page 1 o< 17
                    APPENDIX D

NPSIS SAMPLE RECEIPT SOFTWARE FOR LABORATORIES

-------
                                                            4/5/88
TO:       DATA MANAGER, EFA/TSD LAB

FROM:     CHIP LESTER, ICF IHC.  (

RE:       HPSIS SAMPLE RECEIPT SOFTWARE FOR LABORATORIES
     ICF's National Pesticide Survey Information System (NPSIS) is ready to
collect information from you regarding the receipt of well water samples and
their condition.  Please find enclosed the following items: 1) A users memo
containing all operating instructions, and 2) A copy of Carbon Copy software
which is necessary to establish communications with NPSIS over phone lines.
As mentioned previously, the software allows you to report the receipt of a
one or more sample kits.  It also prompts you for details regarding the
condition of the samples.  Additional features include; a bulletin board whict
allows you to interactively send messages to ICF staff via your computer
keyboard, file transfer, and access to the ICF computerized mail system for
sending memos.  It is also possible for you to speak over the phone to an ICF
staff member during your session.

     It is important that you test the communications link between the NPSIS
computer and yours.  We have experienced trouble when using Carbon Copy
software with a computer which has a Manzana 3.5 inch disk drive, and also
with computers which have a non-Hercules or non-EGA compatible graphics card.

     For testing purposes, your sample kit identification numbers and FedEx
airbill numbers (respectively) are: PD-0000-711 and 1111111111, and PD-9999-
711 and 2222222222.  Use these sample kit identification numbers when trying
out the NPSIS Sample Receipts Program.

     We feel that it would be helpful to both parties if you could call us
when you are ready to test the NPSIS system, and we will assist you over the
phone during your session.  If you would like to do this, please call Beth
Estrada at (703) 934-3431.  NPSIS will be available for access 24-hours a day,
seven days a week.  Ve appreciate hearing any comments you have regarding
NPSIS.

-------
                      THE NPSIS  SAMPLE RECEIPT  PKOGBAM


     NPSIS is designed to keep crack of the day co day operations of the
National Pesticide Survey.   You play an important role in NTS and your timely
notification of receiving a kit of samples is essential to the success of NFS.
Me have designed the Sample Receipt Prograa with your busy schedule in aind.
NFSIS will obtain the minima amount of information necessary while still
maintaining a secure system.  You will be entering data into the NPSIS
personal computer via your own computer,  modem,  and Carbon Copy software.
1.1  Hardware and Software Requirements.
     The NPSIS Sample Receipt Prograa has a minimum hardware and software
requirement.   Here  is a list of items you will need:
          Hardware:
                    One (1) IBM PC,  XT,  AT,  or Personal System model with at
                    least 640K memory.

                    One (1) 2400 or 1200 baud Hayes or Hayes compatible modem
                    with cables. (See Carbon Copy guide for cabling requir-
                    ements and a description of usable modems)

                    One (1) data transmission phone line.
          Software:
               •    NPSIS Sample Receipt Program access provided for you by
                    ICF.

               •    One (1) copy Carbon Copy software which is provided to you
                    by ICF for the duration of NPS.
1^2  Initial Installation Steps.

     Before you can access and use NPSIS. you must first load the Carbon Copy
software onto your PC.   The directions are provided in the Carbon Copy manual.
One item you will want to include is an entry into the "Call Table".  This
entry will include a name, telephone number, and password for the NPSIS
computer.  To enter these items into the Call Table, press "2* from the Carboi
Copy Parameters' Screen.  The information you must enter consists of the
following:

-------
                    Name:   NPS

                    Telephone Number:  703->#1-0629

                    Password:  NPS
1 . 3  Pan
     NPSIS will maintain a set configuration throughout operation.  Any
changes due to updates  in equipment or the system which will affect your
ability to communicate  through Carbon Copy will be forwarded to you.  The
parameters which will be maintained at this time are:

               •    2400 baud modem speed.

               •    Answer ring count equal to one .

               •    Re -boot on exit after 5 minutes.  (If there is a power
                    failure or some other type of interruption, you can log
                    back on to NPSIS and resume your  session.)

               •    Five minute inactivity time constraint.

               •    Two password attempts .
             A SAMPTJt RECEIPT TO HPSIS.
2 1  Establishing a Communications Link
     Once you have installed Carbon Copy and have all of the necessary
hardware, you are ready to "log on" to the NPSIS computer at ICF.  To do this:
          Type: C:> CuiMJ* UPS    in your directory containing Carbon Copy.

This command will automatically dial the NPSIS computer, send your password
for verification, and establish a data link between the two computers.  You
will be able to discern what is taking place by messages to your screen.

                       Receipt Into NPSIS.
     Once you have established a data link, ( e.g., are "logged on"), you will
see on the screen exactly what is on the screen of the NPSIS computer.  This
screen you are viewing is the main menu for the Sample Receipt Program.
Remember that you are controlling the NPSIS computer via a 2400 baud phone
line and your typing will appear on the screen at a much slower rate than you
are accustomed to.  A few tips on how to use the system are outlined in the
next section.

-------
    1  Useful Tins on How to Use  KPSIS.

     Before you start,  a few thing*  to reaember are:

     •    Pressing the "Esc" key  will cancel all changes for the screen you
          are currently in and return you to the previous screen.  Pressing
          •Esc" at the Searching  Screen returns you to the main «enu.

     •    Pressing "PgDn" or "PgUp"  will save the items you have entered in
          the current screen and  place you in the next or previous screen,
          respectively.   This  feature is handy to use when you only have a fe
          items to enter in a  screen which prompts for several items.

     •    Pressing "Enter", "arrow up",  or "arrow down" will move the cursor
          from field to field  in  each screen.  Remember that using the
          sideways arrows will not work.

     •    Pressing the "Alt" and  "Right Shift" keys together will place the
          Carbon Copy Control  Screen over the NPSIS Sample Receipt Program.
          You can then use the communications features in Carbon Copy.
          Pressing "F10" again when  you are through will replace the NPSIS
          Sample Receipt Program  screen you were currently in back on your
          screen, and

     •    Because you will are most  likely to be entering information
          regarding a number of kits at one time, after you save or cancel
          your entries for one kit,  you will be placed at the initial Sample
          Searching Screen for a  new kit.  If you are finished with your data
          entry, simply press  "Esc"  to exit the Sample Searching screen and be
          placed in the main menu.
2.3  A Basic Outline of th« ^*»role Receipt Pr,?Era!Bi

     The NPSIS Sample Receipt Program has three basic features:

               •    Initial reporting of a NPS sample kit of sample bottles.

               •    Ability to edit or re-edit an existing report of a kit
                    receipt,  and

               •    Access to ICFs computerized mail system which provides the
                    ability to send memoranda to ICF staff.

     The information obtained in an entry for a kit of bottles is:

               •    The kit identification number, the FedEx airbill number,
                    and the last name of the person making the entry.
                      A
               •    Any damage to the kit as a whole such as melted ice or any
                    breakage of the cooler.

-------
               •    Verification of which bottles belong in a kit or cooler,
                    notification of any Biasing bottles or any additional
                    bottles, and

               •    Any daaage to each sample bottle which renders it unusabli
                    for analysis and testing.
2 U  NPSTS S«iimle Recetot Program Screens
     When you have completed the logon procedure, you will  see the following
main menu on your computer screen:
                NATIONAL PESTICIDE SURVEY INFORMATION SYSTEM

                SELECTION MENU FOR REPORTING SAMPLE RECEIPTS    04/05/88
                         Report \ Edit a Sample Receipt
                         Send a Memo

                         Press  to Logoff
                        use f land  4-1 to select option.

     The screens provided in this memo will show all of the screens available
and thus represent the maximum number of screens you will encounter with
NPSIS.  It is most likely that you will not have the need to enter informacio
reporting damaged kits or samples.  Therefore, not all of the screens deplete
below will appear in your normal session.

     If you choose the first item on the menu, "Report \ Edit a Sample
Receipt*, you will then be prompted for the kit identification number and  the
FedEx airbill number Associated with the specified kit.  The screen will
appear like this:

-------
               NFS Sample Receipt Searching Screen
         ** Enter the following itens to access kit information **

         To find the Kit information in KPSIS in the no it complete
         and accurate fashion, please enter the Kit number and the
                         FedEx airbill number.

          Enter kit number:
         	> PD-0001-151

          Enter FedEx airbill *:
         	> 1111111111

          Enter your last name:
         	> CHIANG
                  * Press ESC to exit the searching *
     If the kit number you have entered is incorrect, or if the kit number  ant
FedEx airbill number combination is incorrect, NPSIS will prompt you to  try tt
enter these number again, as illustrated on the next page.  It is possible
that the FedEx airbill number on the kit is not the same as the FedEx airbill
number which was entered into the NPSIS system.  This could happen if the
field team loses or damages the airbill.

-------
       ERROR!!   The  kit  you entered cannot be found. .  .

                 Kit number:   PD-0001-151
                                    AND
       FedEx airbill number:   1111111111

            Please check these numbers and try again!

       ******************* * *•* ***************** A********AAA*AA**********
        NPSIS  is designed to track Kits and FedEx airbill numbers.
        The Kit and  FedEx airbill number combination you have entered
        does not aatch what is currently in the system.  Please enter
        the correct  combination.   If you still have problems, try
        leaving the  FedEx airbill * BLANK.  Only enter the Kit number.
       ***********************************************"*****************
Press any key to continue.. .
     Then,  you will  encounter this screen insuring that you have entered the
FedEx airbill number:
    Kit No.:    PD-0001-151
        Did you enter  the  correct Kit number and FedEx airbill number?
        NPSIS is  designed to store and track all FedEx airbill numbers.
        This Kit  may have a different FedEx airbill number than the
        system, please  enter the new FedEx airbill number:
     Note:  if the  correct airbill number was entered before, hit ENTER.
            PgDn (Next page),  PgUp (Previous page), Esc (Exit)

-------
     Once you have correctly Identified the sample kit, NPSIS will ask you
there is any daaage to the kit a* a whole:
   Kit No.:    PD-0001-151


      Was there any danage  to the saople kit?  (Y/N)
               PgDn (Next page),  PgUp (Previous page), Esc (Exit)

-------
     If you press  "Y", NPSIS will then prompt you for  the apparent  cause  of
damage:
   Kit No.:   PD-0001-151


      Was  there any damage to the saaple kit?   (Y/N)


      Please  indicate the cause for damage:

           Kit is broken  (Y/N)  Y

           Ice is melted  (Y/N)

           Other Reason   (Y/N)

       Please enter any comments about the sample kit.

     Comments:  Broken upon arrival.
     Comments:
     PgDn (Next page), PgUp (Previous p«ge), use f ^or  •*—' to select  field.

     There may already be comments regarding the kit  in the comment field
shown in the above screen.  In this case, please enter your comments after any
which already appear.  This insures that no information is destroyed.

-------
     Next,  NPSIS will ask you Co survey the contents of the kit and check tha
which bottles are contained within the kit.  You should then look at the
bottle labels and determine if any are missing.  Don't forget to check and
determine if any bottles have been included in the kit which do not appear on
the list provided by NPSIS on this screen:
   Kit No.:    PD-0001-151
      Please compare the following bottle numbers
            with those in the sample kit.

             Bottle No:    PD-0001-1-1-01
             Bottle No:    PD-0001-1-1-03
             Bottle No:    PD-0001-1-3-01
             Bottle No:    PD-0001-1-3-03
             Bottle No:    PD-0001-1-9-01
             Bottle No:    PD-0001-1-9-03
      Did you receive exactly these bottles in the sample kit?  (Y/N) H
               PgDn (Next page). PgUp (Previous page), Esc (Exit)

-------
     If you have pressed "N", indicating chat you did not receive  exactly what
NFS IS assuaes you have received, you vill ba prompted Co enter  Che appropriate
information.  This information includes pressing a "Y" or a  "N" beside  each
bottle, and entering Che botcle number found on the labels of any  additional
bottles you have received:
   Kit No.:   PD-0001-151

         Please indicate which bottles you received:

            Boccle No:           Received (Y/N)
          PD-0001-1-1-01
          PD-0001-1-1-03
          PD-0001-1-3-01
          PD-0001-1-3-03
          PD-0001-1-9-01
          PD-0001-1-9-03
                              H
                              H
                              Y
                              Y
                              Y
                              Y
    1.
    3.
    5.
    7.
    Please indicate any additional bottles you received:
Bottle No.:  PD-0002-1-1-05   2.  Bottle No.: PD-0002-2-2-01
Bottle No.:  FD-0004-4-4-01   4.  Bottle No.:   -    - - -
Bottle No.;    -    ...     6.  Bottle No.:   -    ...
Botcle No.:    -    ...     8.  Bottle No.:   -    ...
     PgDn (Next page), PgUp (Previous page), use f |or ••—'to select field.

     Notice that the user has indicated chat he did not receive the first two
bottles on the list.  Also note chat the user has indicated additional bottles
which have come in the sample kit, but which were not on  the list.

-------
     N«xc.  NPSIS pronpts you to indicate if any of the individual bottles have
been daaaged and rendered unusable for analysis:
   Kit No.:    PD-0001-151


      Was there any daaage to the «aaple Bottles?  (Y/N) T
               PgDn (Next page),  PgUp (Previous page), Esc  (Exit)

-------
     In order to complete the appropriate information on damaged samples, you
must first press a "Y" or a "N" in the field labeled "Damaged Y/N".  If you
have entered a "Y* in this field, you oust then identify what the cause of the
damage is, to the best of your abilities.  As noted on the computer screen
below, che "Other" category should be used if the sample is unusable but is
not broken.  Please try to connent whenever possible.
   Kit No.:    PD-0001-151

           Please indicate which bottles are damaged by entering Y or N,
           and for those which are damaged, indicate the cause of damage.
                                   •--CAUSE ---
    Bottle No:   Damaged           Broken      Other         Comment
    	    (Y/N)             (Y/N)      (Y/N)

   PD-0001-1-3-01   H
   PD-0001-1-3-03   H
   PD-0001-1-9-01   H
   PD-0001-1-9-03   IT
   PD-0002-1-1-05   H
   PD-0002-2-2-01   Y                Y
   PD-0004-4-4-01   H
   The 'Other'  cause category is for reporting contamination of a sample,
   e.g.  contamination noted on the Sample Tracking Form, air bubbles,
   or other reasons a sample is unusable.
     PgDn (Next page), PgUp (Previous page), use 4 | or *-J to select field

-------
     Now you have completed *11 of the necessary information needed to verify
chat che proper samples have reached their final destination in usable
condition.  You aay save your kit entry by pressing  "Enter".  If you wish to
cancel your kit entry and try again,  press "N" and "Enter".  If you wish to
view or edit the current kit entry,  press "R" and "Enter" and HPSIS will plac
you back at the beginning of your entry.
     You have completed all of the data entry screens for this Kit.

     You may save your entry by pressing 'Enter' .

     You may cancel your entry by pressing 'N'  and 'Enter'.

     You may verify or edit this entry by pressing 'R' and 'Enter'.


                    * * *  Accept entries?  * * *
                * Press    *—      to Save           *
                * Press N and •*•— '  to Cancel         *
                * Press R and ^-J  to Verify or Edit * Y
     By pressing "Enter" ,  you have saved all of the information necessary for
a particular sample kit.  NFS IS assumes that you will enter more than one kit
entry per session.  Therefore, you will be placed at the initial "Searching
Screen".   If you are finished, press "Esc" and you will be returned to the
main menu.  You can then log off of NPSIS by pressing "Alt" and "Right shift"
at the same time.  You may also send a memo through the ICF computerized mail
system.  To do this, cursor down to the second menu choice and press "Enter".

     The next two pages of this memo describe how to use the ICF electronic
mail system.  Note that the password for you is NFS.  The mail system software
program will prompt you for this password before it will allow access to the
system.  Also, when you are selecting the recipients of your memo, please
press the space bar beside the initials "NFS".  This will send your memo to
all ICF staff involved in the NFS project.  If you wish to send memos to a
particular ICF staff member, please call Beth Estrada for the identification
number of the desired ICF employee.

-------
                                  ELECTRONIC MAIL
Function
Augment   office   communications
electronic transfer of notes and files.
with
Summary
Electronic Mail (E-Mail) allows you to send,
receive,  read,  and  subsequently  save  or
discard notes and attached files.

When  you  power  up your workstation  you
will automatically enter E-Mail if  you have
received any mail.   Enter your password to
check  your  mail,  or press   twice to
avoid  E-Mail and  continue  to  the Assist
main menu.
Instructions
Operation  of E-Mail  is  similar to  Lotus
1-2-3.  Press the Fl key  to receive help at
any  time during  operation.   If  any more
help is needed contact workstation support
to receive a manual.

For  more  information on  any  feature  of
electronic mail, use Network Courier's  on-
line help or refer to the User's Manual.
Passwords

Your password will be "password* until  you
change  it yourself.   Once  you have  given
your password  and entered  E-Mail, you  can
change  your  password by selecting Options,
then Password.
Reading Mail

 1.    Select   "Read"  from  your  menu
      Highlight    read,then    press
      .
 2.    Select the note to read:
      a.    Highlight the note (using
           the arrow keys); and press
           .
      B.    To save the note, select
           •Storage", then "Save". Enter
           the name of the file to which
           the note should be saved.
 3.    Press  to select another note.
        Writing Mail

         1.    Select "Compose", then "edit".
         2.    Press  when the highlight
              moves to TO".
         3.    Select the recipients(s):
              a.   Move the  highlight to the
                   first recipient's initials.
              b.   Press the space bar. A
                   small mark will appear.
              c.   Repeat steps a and b for all
                   recipients. Press the space bar
                   twice to "de-select" recipients.
                   The small mark will disappear.
              d.   Press <£SC> to cancel the
                   entire list.
         4.    Select the initials of those who will
              receive copies:
              a.   Press the  down arrow to  mo%
                   to "CC".
              b.   Select recipients as instructed
                   above (step 3, a-d)

-------
Writiaf Mail, continued

  5.    Enter t subject and priority.
       (optional)
  6.    Select attachments (optional):
       a.    Press <£NT£R> and type the
            path for the document(s).
       b.    Press  and select the
            document(s) to be attached.
       c.    Repeat steps a and b for
            documents in another directory.
  7.    Enter the text of your message.
  8.    Press  when  finished.
  9.    Select Transmit" to post the note
       and attachments.

Quitting th* Mail Program

  1.    Press  from the menu.
  2.    Select "YES".

-------
                          Appendix E
                          Revision No. 2
                          Date: July 29, 1990
                          Page 1 of 6
      APPENDIX E

FORMAT FOR NPS DATA

-------
               FORMAT FDR NATIONM, PfcSTICTrgr SURVEY (NFS) DATA
LBffi   COOMB

 1       1-6       l.Tenp
         9-14      S_Tenp
        17-24      Date_Sam
        27-34      Date_Shp
        37-44      Date_Rec
        47-54      Time_Sam
        57-64      Time_Ice
           [FOR MEMDS 5 AND 9 ONLY]
        68-69      pH

 2       1-6       alter INITIAL TEMPHRA3URE OF
         9-14      enter SIABHIZED TEMPEKAIURE CF
        17-24      enter EME SftMFLEE
        27-34      enter EME atTPPED
        37-44      enter EftlE RhnT'.ivtu
        47-54      enter TIME SAMPLED
        57-64      enter TIME ICED
           [FOR METHDS 5 AND 9 CKLY]
        67-70      enter pH

 3      BLANK

 4       1-17      Receipt Condition

 5       1-80      enter CONDITION OF  SAMPLE UPON RECEIPT AT LABOFATORY

 6      BLANK

 7       1-6       Sanp #
        16-18      Lab
        21-25      Set #
        28-35      Date_Spk
        38-45      Date_Ext
        48-55      Date.Ana
        58-63      Colum

 8       1-13      enter SAMPLE IDENTIFICATION NCM3ER
        16-18      enter LAB ABBREVIATION
        21-25      enter SET NIMER
        28-35      enter DATE SPIKED
        38-45      enter DATE EXTRACTED
        48-55      enter DATE ANALYZED
        58-63      enter ANALYSIS COLUM

  9     BLANK

-------
           FCWftT FCR NATIClttL PES'l'lClUE SURVEY (NFS)  EKTA (cent.
LEE
 10      1-4       Type
         8-13      spiker
        16-22      Detract
        2S-31      Analyst
        34-40      San\_Vbl
        43-49      Bct_Vt>l
        52-60      Int.  Std.
        65-70      % Surr

 11      1-5       enter SAMPLE TYPE
         8-13      enter SPIKER'S INITIALS
        16-22      enter EXTRACTOR'S INITIALS
        25-31      enter ANALYST'S INITIALS
        34-40      enter \^LUME CF S>WFLE
        43-49      enter VOLUME CF EXTRACT
        52-62      enter INTETOPkL SIMJEftRD
        65-70      enter PERCENT RECCVERY CF SURROGATE

 12     BLANK

 13      1-8       Connents

 14      1-80      enter ANY PERTINENT CdMENTS ON SAMPLE AND ANALYSIS

 15     BLANK

 16      1-7       Analyte
        29-33      Cone.
        39-45      Analyte
        67-71      Cone.

17-?     1-25      enter ANALYTE 'S NAME
        28-34      enter CCWCENIRATICN OR PERCENT RECOVERY
        39-63      enter ANALYTE»S NAME
        66-72      enter CCNZENTRATICN CR PERCENT RECOVERY

-------
      PQRAT FOR NATIONAL ttifl'lClUE SURVEY (NFS) HJSIKMNr CONTROL DMA
T.TNTE   CDLIMB
 1       1-3       Lab
         6-11      Method
        14-21      Date_Ana
        24-30      Analyst
        35-37      S/N
        42-44      PSF
        49-51      Pty
        55-58      Res.

 2      BLANK

3-?      1-3       enter LAB ABBREVI7MICN
         6-11      enter MEUHDD NUTBE3*
        14-21      enter DAIE ATftLYZED
        24-30      enter ANALYST'S INITIALS
        33-37      enter SIOftL TO NOISE RATIO
        40-44      enter PENH SYMMETRY FACTOR
        47-51      enter FfWC GEOMETRY FACTOR
        54-58      enter RESOLUTION

-------
                         NOTES CN NFS EftlA FCEMVTS



1. The format for any date is mm/dd/yy

      A missing date should be entered 01/01/60

2. The format for any time is hh:mm in military time

      A missing time should be entered 00:00

3. Any other data that is missing should be entered with a period  (.)

4. The number of decimal places should be as follows:

      Concentration          3
      Percent Recovery       1
      Internal Standard      0
      Instrument Controls    2
      PH                     1
      Temperatures           0
      Volumes                0

5. The codes for Column are as follows:

      Primary          PRIM
      Confirmatory     OUT
      Third            QCMS

6. The codes for Lab are as follows:

      TSD                                       TSD
      CFP                                       GPP
      WERL                                      WER
      Radian                                    RAD
      Battelle                                  BCD
      James M. Montgomery                       JMM
      Alliance                                  ALL
      Environmental Sciences and Engineering


7. The codes for Type are as follows:

      Field Sample                 SWF
      Shipping Blank               SBLK
      Method Blank                 «LK
      Lab control Standard         LCSt
      Lab Spike Sample             LSStt
      Time Storage for Extract     H1BP
      Tune Storage for Sample      KT9P
   where § is the mixletter (A,B,C or D)
    and # is the spiXing level (1,2 or 3)
                     \\

-------
                      NOTES CN NFS IMA FCE9AX5 (CCRt.)
 8.  There should be at least cne blank line between samples in the NFS data
    file.

 9.  The codes for Concentrations and Percent Recoveries are as follows:

       Not Analyzed
       Not Detected (< Estimated Detection Limit)            -999
       Saturated                                            -777
       Other                                                -333
       Below Reporting Limit, but above HX                  -ill

       Above Reporting Limit, but not Quantified              888

10.  If a reported value is greater than (>) some number in the NFS instrument
    control data, then use a minus sign (-) instead of >

-------
                              Appendix F
                              Revision No. 2
                              Date:  July 29, 1990
                              Page 1 of 3
            APPENDIX F

DATA REPORTING FORMAT CHANGES

-------
DATS:    April 18, 1988

SUBJECT: Data Reporting Format Changes

FROM:    David J. Hunch, Chemist
         Drinking Water Quality Assessment Branch

TO:      UPS Technical Monitors (See below)
    The purpose of this memorandum is to consolidate the changes to the
UPS data reporting format, which have occurred since it was originally
constructed. Too have previously been supplied with most of these changes,
but please check to be sure that they have all been relayed to your
contract and referee laboratories.

    1.   Line 2, columns 1-6 are to be used to record the pH measured in
         the field. This data will be found on the field sample tracking
         sheet.

    2.   Line 2, columns 67-70 are to be used to record the pH measured
         upon sample receipt at the laboratory. This only applies to
         methods 5 and 9.

    3.   Line 8, columns 1-13, Sample Identification lumber, have been
         expanded to columns 1-14.

    4.   The data entered on line 10, coloans 52-60, concerning the
         internal standard, it should be entered not as the peak area but
         as the "percent recovery" as compared to the mean observed for
         the calibration curve.

    In order to simplify the "Sample Type* code  (line 11, columns 1-5),
the following codes should be used to designate the various types of
spiked samples.
    LCSt
    LSStf
    DTSt
    BTKt
    BTSt
Laboratory Control Sample
Laboratory Spiked Sample
Day 0 Time Storage Sample
Extract Time Storage Sample
Sample Time Storage Sample
    In additioa, two clarifications have b*«a made  to  the codes for
analyte concentration entries.

    -999 » let Detected  « 1/2 Minimum 1*sorting Limit)
    -111 • Below Minimum Reporting Limit bet greater than or equal to
           1/2 the Minimum Reporting Limit.

-------
3

4

5

6

7
8
                                Note: Method 9 only
                Foraat for National Pesticide Survey (NFS) Data

        COLUMNS    DESCRIPTION
1-6
9-U
17-24
27-34
37-44
47-54
57-64
68-69

1-6
9-14
17-24
27-34
37-44
47-54
57-64
68-69

BLANK

1-17

1-80

BLANK

1-6
16-18
21-25
28-35
38-45
48-55
58-63

1-13
16-18
21-25
28-35
38-45
48-55
58-63
S'Temp
Date_Saa
Date~Shp
Date'Rec
Ti«e~Sam
Time'lce
pB

enter
enter STABILIZED TEMPERATURE OF VATER
enter DATE SAMPLED
enter DATE SHIPPED
enter DATE RECEIVED
enter TIME SAMPLED
enter TIME ICED
enter pB    Note: Method 9 only
Receipt Condition

enter CONDITION OF SAMPLE UPON RECIEPT AT LABORATORY
Samp t
Lab
Set t
Date Spk
Date'Ext
Date"Ana
Column

enter SAMPLE IDENTIFICATION NUMBER
enter LAB ABBREVIATION (JMM)
enter SET NUMBER
enter DATE SPIKED
enter DATE EXTRACTED
enter DATE ANALYZED
enter ANALYSIS COLUMN
        BLANK

-------
                           Appendix G
                           Revision No. 2
                           Date: July 29, 1990
                           Page 1 of 4
      APPENDIX G

DATA REPORTING CODES

-------
   DftTE: September 9, 1988

SUBJECT: Data Reporting Codes

   FECM: Christopher Frebis, CSC Statistician

     TO: Distribution


     Hie purpose of this memorandum is to discuss the reporting codes used in
the National Pesticide Survey.  There has been sane confusion over these codes
as to when and where to use them and their exact meaning.

     Table 1 identifies the unique sample types (SAMP - field sample, MSLK -
method blank, SBLK - shipping blank, LCS - lab control standard, and LSS, DTS,
HIE, and HTS - spiked field samples — these last three are each a type of time
storage sample).  Under each unique sample type are the only possible codes
that can appear for that sample type.  (Note: -555 has been added for the
situation where the contract lab sends the extract to the referee lab for QCMS
analysis, and the code -222 has been deleted.)   There is also a type of
decision tree for field samples since they are a little more complicated with
three analyses for confirmation and qualitative only analytes.

     I hope this memorandum helps to put everyone on similar terms as well as
clearing the muddy water.  If there are any questions of different scenarios
you wish to discuss, please call me at (513) 569-7498.


Distribution:  Herb Brass, Technical Support Division
               Aubry Dupuy, Environmental Chemistry Laboratory
               Carol Madding, Technical Support Division
               Bob Maxey, Environmental Chemistry Laboratory
               Dave Munch, Technical Support Division
               Kent Sorrell, Technical support Division
               Bob Thomas, Technical Support Division

-------
                      TABLE 1:  USES CF DATA CODES IN NFS

                                 SAMPLE TYPE
                                                                   LSS,DTS
SAMP
,(a)
_ i i i (C)
_333
-444 e
-555 
-666 
_777(h)
888 (V
-999 (j)

MBLK
.(a)
—111
-333 
****
-777(11)
888 (V
-999(1))
PTHv(^)
SBLK
.(a)
_j_^j_(c)
-333(d)
-444(6)
-555 
-666 (9>
-777 (W
888(V
_999(j)
rvM-ir" ikl
LCS
(b)
•
****
****
****
****
****
-777 (h)
****
****
4 rof ( 1 )
fflE,HES
.(b)
****
-333(d)
-444(6)
****
****
_777(h)
****
****
_ __ (1)
(a) Analyte dropped from survey (Daneton-s and Carboxin sulfoxide)  or not
    analyzed on the second colurm or in GC*E analysis.

(b) Analyte not in mix.

(c) Analyte's concentration between MRL/2 and M?L.   (If no confirmation is run,
    a coninent as to why should be made.)

(d) A lab mishap, e.g.  sample lost during extraction or sample dropped etc.,
    •Diis is a unique situation.   (A comment should give further explanation.)

(e) Tliis analyte fails  QC in  this set (e.g.  LCS  out  of  control or positive
    method blank or time to extraction or analysis is too long) and therefore
    cannot be reported,  however the analyte does not require a qualitative
    challenge.  TJiis code also applies to any spike  sample in a set where the
    LCS is out of control.

(f) GCMS only:  Sent to referee lab for GOB analysis.

(g) Inis analyte fails  QC in  this set and therefore  cannot be reported, however
    the analyte requires a qualitative challenge.

(h) Analyte was saturated.  Should be diluted and re-done, if observed in a
    field sample.  (Another sample with the exact same  header information
    should appear, analytes not saturated in the original sample should be
    reported as ., and  saturated analytes should be  reported as their
    concentration.)

(i) Positive, can occur in two fashions: 1) any  analyte in GCMS analysis; or
    2) a qualitative only analyte on either of the first  two columns.

(j) Analyte's concentration below MRL/2.

(k) Concentration above M3L for quantitative analytes,  reported to three
    significant figures.

(1) Percent recovery, reported to one decimal place  (even if recovery is 0.0%).

**** Code not applicable.

-------
                               Saocle (Qualitative cnlv analvte)
                   -333
                      -999
                                888
                                              -333    -999     888
                                                    -333   -
                                                           -999    888
                    GOUMJ

                     FRIM


                     OUT
                                                   -999     888     GCMS
                                                                   GCMS
                                                               (at  referee)
      Samplg  (CV?ntL^t-?'t'iv'?
                                                   with CC
-333
                         -444
                                                         -666
                                                 -333    -666    -999
                                                -333  -555   -999   888
                                                -333  -999    888
                     FRIM


                     CCNF


                     GCMS
                                                                  GCMS
                                                               (at referee)
     Sarrle  (Quantitative
                                                       rp nt
                  -Ill
      -333
                      -777*     -999
                     FRIM
                                                            CTKF
-333 -555 -999 888  -333 -555 -999 888   -333 -555 -999 888  -33'3 -5$5 -999 888   GCMS
333   ~999   c°nc
                                                           core
                                                       -333   -999   core:
-333 -999  888
    -333 -999  088
                   -333 -999  888
-333 -999  888       GCMS
                 (at referee)
      a = Dilute and reanalyze

-------
                             Appendix H
                             Revision No. 2
                             Date: July 29, 1990
                             Page 1 of 11
         APPENDIX H

NPS RAPID REPORTING SYSTEM

-------

   i
   |    UNITED STATES ENVIRONMENTAL  PROTECTION AGENCY
                           CINCINNATI. OHIO 45268
MEMORANDUM

DATE:    April 12,  1988

SUBJECT:  NPS Rapid Reporting System
FROM:    David J. Munch, Chemist
         Drinking Vater Quality Assssment Branch

TO:      NPS Technical Monitors
    Jerry Kotas has requested that any confirmed results of health
significance be reported as quickly as possible.  Therefore, if an analyte
listed in the attached tables is observed in the primary analyses, at or
above the rapid reporting limit, the following actions should be
instituted.  For any listed analyte where the rapid reporting level is
less than or equal to 1/2 the minimum reporting level  (MRL), any
occurrence at or above 1/2 the MRL should also be processed as below.
(Note:  The procedures for determining the occurrence of NPS analytes that
may occur below the MRL, and are not listed on the attached tables, have
not yet been finalized.)

    1.   The appropriate confirmational analyses (CC/MS for methods 1-3,
         6-7, second column for Method 5) should be performed as soon as
         practical.

    2.   The laboratory should telephone their Technical Monitor, the same
         day the confirmation is completed.

    3.   The laboratory should immediately document the observed result in
         a letter to their Technical Monitor.

    4.   As quickly as possible on the day the above telephone call is
         received from the laboratory, the Technical Monitor should inform
         their Laboratory Analytical Coordinator of the finding.  The
         Technical Monitor should forward on to the Laboratory Analytical
         Coordinator the above documentation, with any comments he/she may
         have concerning the validity of the result.

    5.   The Laboratory Analytical Coordinator should inform Jerry Kotas
         and the second Analytical Coordinator of the finding by telephone
         the same day if possible, and in writing after the documentation
         is received from the Technical Monitor.

    6.   The Analytical Coordinators are to request, through the
         appropriate Technical Monitors, that all analyses  for this sample
         site be conducted, and reported in writing, as soon as practical.

-------
                                    -2-
    If you have any questions concerning these procedures, please let Bob
Maxey or me know.  Also, please pass on this information to your contract
and referee laboratories.  They will need to have this information in hand
prior to their conducting the dry run.

Attachment

Addressees:

    A. Dupuy
    L. Kamphake
    C. Madding
    R. Maxey
    R. Sorrell
    R. Thomas
cc:
    J. Kotas
    H. Brass
    A. Kroner
    J. Orme

-------
                 METHOD ffl
 AMALYTI                  RAPID REPORTING  LEVEL

Alacblor                           44  Ug/L
Aaetryn                           300  ug/L
Atrazine                           35  ug/L
Broaacil                       2,500  ug/L
Butylate                          700  ug/L
Ctrboxio                       1,000  ug/L
Diphenaaid                        300  ug/L
Fenaaiphos                        5.0  ug/L
Hexazinone                     1,050  ug/L
Hetolachlor                       300  ug/L
Hetribuzin                        250  ug/L
Propazine                         500  ug/L
Siaazine                           50  ug/L
Tebuthiuron                       125  ug/L
Terbacil                          250  ug/L

-------
                 METHOD 12
   AHALYTE                    RAPID REPORTING LEVEL

alpha-Chlordane                     0.5 ug/L
gamna-Chlordane                     0.5 ug/L
Cblorothalonil                      150 ug/L
Oactbal (DCPA)                    5.000 ug/L
Oieldrin                            0.5 ug/L
Propacblor                          130 ug/L
Trifluralin                          25 ug/L

-------
                 METHOD 13
   AMALYTE                RAPID REPORTING LTYEL

Acifluorfen                      130 ug/L
Bentazon                        87.5 Uff/L
2,4-D                            100 ug/L
Dalapon                          800 ug/L
Dicamba                           13 ug/L
Dinoseb                          3.5 Ug/L
Pentachlorophenol                300 ug/L
Picloram                         700 ug/L
2,4,5-T                          105 ug/L
2,4,5-TP                          70 ug/L

-------
              METHOD f4
  ANALYTE               EAPID REPORTING LEVEL

Cyanazine                     13 ug/L
Diuron                        70 ug/L
Fluometuron                  438 ug/L
Prophan                      595 ug/L

-------
              METHOD 15
 AMALYTE                RAPID REPORTING LEVEL

Aldicarb                       10 ug/L
Baygon                         40 ug/L
Ctrbtryl                    1,000 ug/L
Carbofuran                    50 ug/L
Hethomyl                      250 ug/L
Oxamyl                        175 ug/L

-------
                 KETEOD §6






    XMALYTI                RAPID REPORTING LEVEL




ethylene thiourea               l.OS ug/L

-------
                           METHOD 17
	AJfXLYTE	          RAPID REPORTING LEVEL

dibroaochloropropane                          2.5 ug/L
1,2-dichloropropane                            56 ug/L
cis/trans 1,3-dichloropropene                  11 ug/L
ethylene dibromide                           0.04 ug/L

-------
            METHOD 19





    AMXLYTI            UPID IPOtTIMG



Nitrate/Nitrite        10,000 ug/L

-------
DIXON'S TEST

Dixon's test is used to confirm the suspicion of outliers of a set of data
(for example, control chart data points).  It is based on ranking the data
points and testing the extreme values for credibility.  Dixon's test is based
on the ratios of differences between observations and does not involve the
calculation of standard deviations.

The procedure for Dixon's test is as follows (from Taylor, 1987):

     1)   The data is ranked in order of  increasing numerical value.  For
         example:
     2)   Decide whether the smallest, X^  or the largest, X^,  is
         suspected to be an outlier.

     3)   Select the risk you are willing to take for false rejection.
         For use in this QAPP we will be using a 5% risk of false
         rejection.

     4)   Compute one of the ratios in Table 1.  For use in this QAPP we
         will be using ratio r22,  since  we will  be  using between 20  and
         17 points for the control charts.

     5)   Compare the ratio calculated in Step 4 with the appropriate
         values in Table 2.  If the calculated ratio is greater than the
         tabulated value, rejection may be made with the tbulated risk.
         Fort his QAPP we will be using the 5% risk values (bolded)  .

Example (from Taylor)

     Given  the  following  set of  ranked data:

         10.45, 10.47, 10.47,  10.48, 10.49,  10.50,  10.50, 10.53, 10.58

     The value  10.58 is suspected  of being an  outlier.

     1)   Calculate ru

                       10.58  -  10.53      0.05
              rn  -    .............   -   ----    - 0.454
                       10.58  -  10.47      0.11

     2)   A 5% risk of false rejection (Table 2), ru - 0.477

     3)   Therefore there is no reason to reject  the value 10.58.

     4)   Note that at a 10% risk of false rejection rn  - 0.409, and the value
         10.58 would be rejected.

-------
                                    TABLE 1

                             CALCULATION OF RATIOS
Ratio
f. „
For use if
n is between

if Xn is
suspect
f V V \
(Xn ' Xn-i)
if X: is
suspect
/Y V \
\&2 ~ *-l)
Note that  for use in this  QAPjP ratio r22 will be used.
                                                   V   \            / V
                                                   "n-1/            \A2  ~
                           8  -  10            	         	-
                                                  -  X2)
                                                   V   \            / V
                                                   •"Ti-2^            ^A3
        c2i                11 - 13            	         	
                                                  -  x2)           (Vi
                                                 • Xn.2)            (X3  - Xj]
        r22                14-25            	         	
                                                  '  X3)            (Xn-2 ' Xj

-------
     •10
     •11
     L21
                                    TABLE 2
               VALUES FOR USE WITH  THE DIXON TEST FOR OUTLIERS

                                         Risk of False Rejection
                   n          0^5%           1%            5%           10%
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
0.994
0.926
0.821
0.740
0.080
0.725
0.677
0.639
0.713
0.675
0.649
0.674
0.647
0.624
0.605
0.589
0.575
0.562





0.988
0.889
0.780
0.698
0.637
0.683
0.635
0.597
0.679
0.642
0.615
0.641
0.616
0.595
0.577
0.561
0.547
0.535
0.524
0.514
0.505
0.497
0.489
0.941
0.765
0.642
0.560
0.507
0.554
0.512
0.477
0.576
0.546
0.521
0.546
0.525
0.507
0.490
0.475
0.462
0.450
0.440
0.430
0.421
0.413
0.406
0.806
0.679
0.557
0.482
0.434
0.479
0.441
0.409
0.517
0.490
0.467
0.492
0.472
0.454
0.438
0.424
0.412
0.401
0.391
0.382
0.374
0.367
0.360
Note that for this QAPjP  the  5%  risk level will be used for ratio r
                                                                    22-

-------
Reference:
     John K. Taylor,  Quality Assurance  of Chemical Measurements. Lewis
     Publishers,  Chelsea, MI, 1987.
               U.S. Environmental Protection Agency
               Region 5, Library (PL-12J)
               77 West Jackson Boulevard, 12th Floor
               Chicago, IL  60604-3590

-------
                             Appendix J
                             Revision No. 2
                             Date:  July 29, 1990
                             Page 1 of 2
           APPENDIX J
STANDARD OPERATING PROCEDURE:
      BATCH FILING SYSTEM

-------
                                                             Appendix J
                                                             Revision No. 2
                                                             Date: July 29, 1990
                                                             Page 2 of 2
                                         APPENDIX J
                             STANDARD OPERATING PROCEDURE:
                                    BATCH FILING SYSTEM
 PURPOSE
         To centralize the storage of ESE analytical data batches and all associated documentation for
 each batch.

 PROCEDURE
         All data batches created by the various departments will be placed in departmentally
 assigned colored batch files.  All associated documentation for each batch will be included in each
 folder as well as a documentation checklist. The checklist will be marked by the analyst noting
 everything included in the batch file and will be signed and dated by the analyst and a review person.
         The batch file must also have the "Computer QC Checks" and the "Internal QA/QC Batch
 Checklist" section located at the end of each batch marked, signed and dated by the analyst.  The
 batch then must be reviewed, signed and dated by the Department Manager. If the batch fails any of
 these checks, the corresponding Lab Coordinator(s) for all samples in the batch must also sign and
 date the batch and may add comments.
         The batch file is then signed-in to Information Services to document chain-of-custody of the
 raw data.
         Each batch will then be finalized and filed numerically by department in locked file cabinets
 located in the Information Services department.  Each department manager will  have a key to his/her
 file cabinet and Information Services will retain a key to all cabinets.
         Since only the  most recent batches can be filed in the file cabinets, all "older batches are
 filed by department in a separate, locked storage room with access available only to Information
 Services.
         All data batches including those in the storage room are available for checkout at any time.
 All batches are signed out to the individual with  a hard copy as well as an electronic file kept of all
 checkouts.
U.S. Environmental Protection Agency
Region 5, Library (PL-12J)
77 West Jackson Boulevard, 12th Floor
Chicago, IL   60604-3590

-------