xvEPA
             United States
             Environmental Protection
             Agency
             Effluent Guidelines Division
             WH-552
             Washington, DC 20460
EPA 440/1-79/102
July 1979
             Water and Waste Management
Self-Monitoring Program
Analytical Methods Package
             (Organic Chemicals)

-------
      Self-Monitoring Program
     Analytical  Methods Package
U.S. Environmental  Protection Agency
    Effluent Guidelines Division
      Organic Chemicals Branch
           July 19/9

-------
                             Program Notice
At the time of printing of this package a list of the specific
priority pollutants discharged by each product/process for which
self-monitoring would be required was not available.  However,
based on EPA's experience in the verification program, it can be
assumed that a maximum of 4 methods for organic pollutants would
have to be used.  The methods would most likely be used for the
analysis of volatile organics (2 methods), phenols, polynuclear
aromatics.  For the metals which have been found in organic chemicals
wastewaters, atomic absorption spectroscopy would be used.

The self-monitoring program is intended to provide long-term data
on only those priority pollutants present in wastewaters from the
product/processes or entering the treatment system.  Continued
analysis for the entire spectrum of 129 priority pollutants throughout
the course of the 30 day sampling will not be required.

-------
                                 Index
Chapter 1
     Analytical Methods for the Verification
       Phase of the BAT Review                       Edited by
       (GC)                                          William F. Cowen
                                                     Catalytic

Chapter 2
     Quality Assurance Program for the
       Organic Chemicals Industry (Self-Monitoring
       Phase)

Chapter 3
     Proposed Priority Pollutant
       Test Procedures                               U.S. EPA
         (GC:601-613) page 1                         Environmental  Monitoring &
         (GC/MS:624-625) page 205                      Support Laboratory
                                                     Cincinnati, Ohio  45268

Chapter 4
     Quality Assurance and Quality Control
       Procedures for Screening and Verification
       of Industrial Effluents for Priority
       Pollutants - EMSL

Chapter 5
     Cost Estimate for Self-Monitoring
       Program

-------
      Chapter I
    ANALYTICAL METHODS FOR THE

     VERIFICATION PHASE

      OF THE BAT REVIEW
M>v *-«^ »f» J» '-• §«••«:•» »••*»••» •«• •jr' <****i «r» "
PRELiMiNARi:
 REVISED
March 21, 1979

-------
                       INDEX OF ANALYTICAL METHODS
Procedure Code No.                             Title
         1             Direct Aqueous Injection Procedure for GC Analysis
                       of Acrolein and Acrylonitrile

         2             Method for Benzidine and Its Salts in Wastewater

         3             Method for Organochlorine Pesticides and Phthalate
                       Esters In Industrial Effluents

         A             Total Cyanide

         5             A-26 Resin/GC-FID Method for Phenolr

         6             Analysis of Nitrosamines

         7             Microextraction Method for Organic Compounds in
                       Industrial Effluents

         8             Purge and Trap Procedures for Analysis of Volatile
                       Organic Compounds in Effluents

         9             Method for Polychlorinated Biphenyls (PCBs) in
                       Industrial Effluents

        10             Analysis of Arsenic and Selenium in Industrial
                       Effluents by Flameless Atomic Absorption Spectrophotometry

        11             Analysis of Silver, Antimony, and Thallium in Industrial
                       Effluents by Flameless Atomic Absorption

        12             Analysis of Beryllium, Cadmium, Chromium, Copper, Nickel,
                       Lead, and Zinc in Industrial Effluents by Flame or
                       Flameless Atomic Absorption

        13             Mercury in Water (Manual Cold Vapor-Atomic Absorption
                       Technique)

        14             Pentane Extraction of Organics in Wastewaters for
                       GC Analysis

        15             Acid Extraction Procedure for Phenols

        16             Analysis of Nitroaromatics

        17             Analysis of Polynuclear Aromatic Hydrocarbons in
                       Industrial Wastewater
                              4

-------
Procedure Code:  1

Variations:   1-3
    REVISED
Title:    Direct Aqueous Injection Procedure for GC Analysis      ~;
          of Acrolein and Acrylonitrile

1.   Summary

          This method is similar  to that given by ASTM  (1), method designation
     D3371-74T, "Tentative Method of Test for Nitriles  in Aqueous Solution by
     Gas-Liquid Chromatography."  The detection limit is approximately 1 rag/1
     with this method.  For increased sensitivity,  the  heated purge-and-trap
     procedure (Procedure Code #8) is recommended.

2.   Sampling and Analysis

          No sample extraction or cleanup procedures are employed in  this
     analysis.  Samples should be collected in Teflon-septa sealed-glass vials
     headspace-free.  A small volume (1-20 yl) of aqueous sample is injected
     directly into the gas chromatograph for analysis on a porous polymer
     column, which elutes the water in the sample ahead of the analyte peaks.

3.   Quality Control

          Duplicate and spiked sample analyses are  recommended as quality
     control checks.  A standard  of 1 mg/1 in water is  injected (1-20 yl) ontc
     the GC column to provide an  external standard  for quantification of
     samples.

4.   Variations to Procedure Code #1
     Variation
        No.
Effective
  Date

June,  1977
Description

GC Column:  Chromosorb 102, 60/80
mesh, in a  2 meter x 1/8" O.D.
stainless steel column.

Carrier: He at 30 ml/min.

Injector:  250°C

Detector:  FID at 200°C

Temperature:  100 C isothermal
(Source: Envirodyne Engineers,  Inc.
         for Acrylonitrile)

-------
     Variation      Effective
        No.           Date).  .            Description

        2           March,  1979       ~'GC Column:   Chromosorb 101,"
                                                    60/80 mesh, in a 61 x 4 mm
                                                    ID glass column

                                        Carrier:   He at 60 nl/min. -'

                                        Injector:   200°C

                                        Detector:   150°C, FID

                                        Temperature:  120°C isothermal
                                        (Source:   Southwest Research Institute,
                                                  for Acrolein)

        3           June, 1977          GC Column:   Chromosorb 101,  in a
                                                    6' x 2imn ID glass column

                                        Temperature:  135  isothermal

                                        Injector:   225°C

                                        Detector:   FID

                                        Carrier:   He at 25 ml/min.
                                        (Source:   Envirodyne Engineers)

5.   References

     I.   1976 Annual Book of ASTM Standards (Water), Part 31, ASTM, Philadelphia
          Pa., pp 534-536 (1976).

-------
           Proposed pH7 Buffer for Acrylonitrile-Acrolein Samples


Design Buffer Index:^8 * 5X10   mole/1 (after dilution,
      (dc  \ _ a      1 ml buffer + 40 ml sample)
     d pH / s«/

Calculated Buffer Concentration after dilution with sample:  0.091H,

Est. Ionic strength from buffer * 0.2

Est. /HPO* -  0.28
Est.   H2PO~ -  0.73

pK°a2  -  7.198 for H3P04

Calculated species composition after dilution:  0.034M KH.PO ;  FW » 136.1

                                                0.057M Na2HP04; FW - 141.9

Preparation;

Weigh out 18.6g anhydrous KH PO  and 32.2g anhydrous Na.HPO,.   Pour the salts

Into a 100 ml volumetric flask  and dilute to volume with organic-free distilled

water.

Use;

Add 1 ml of buffer to each clean 40 ml sample vial before taking the vial into

the field.  At the plant site, pour the sample into the vial,  cap the vial,  and

invert for a short time to mix the buffer with the sample.   Check the pH of  the

sample at the time of analysis and record its value.
                                                 Bill Cowen
                                                 May 1,  1979
                                 J--J

-------
Procedure Code:  2

Variations:  1

Title:  Method for Benzidine and Its Salts in Wastewater*

1.   Scope and Application

     1.1  This method covers the determination of benzidine and its salts in
          wastewaters.  The method can be modified to apply also to the
          determination of closely-related materials, as described under
          Interferences (4.2)

     1.2  The salts of benzidine, such as benzidine sulfate, are measured and
          reported as benzidine.

     1.3  The method detection limit is 0.2 yg/1 when analyzing 1 liter of
          sample.

2.   Summary

     2.1  The water sample is made basic and the benzidine is extracted with
          ethyl acetate.  Cleanup is accomplished by extracting the benzidine
          from the ethyl acetate with hydrochloric acid.  Chloramine T is
          added to the acid solution to oxidize the benzidine.  The yellow
          oxidation product is extracted with ethyl acetate and measured with
          a scanning spectrophotometer.  The spectrum from 510 nm to 370 nm is
          used for qualitative identification.

3.   Hazards

     3.1  Benzidine is a known carcinogen.  All manipulations of this method
          should be carried out in a hood, with protection provided for the
          hands and arms of the analyst.  Consult OSHA regulations (1) before
          working with benzidine.

4.   Interferences

     4.1  The multiple extractions effectively limit the interferences to
          organic bases.  The oxidation with Chloramine T to form a yellow
          product is very selective and has been described in detail (2,3).
          The use of the absorption spectrum for the identification of
          benzidine results in a highly specific procedure.

     4.2  Some compounds having a structure very similar to benzidine will
          interfere with the quantification, if present.  Examples of these
          interfering compounds are dichlorobenzidine, o-tolidine, and
          dianisidine.
*From EPA Environmental Monitoring and Support Laboratory, Cincinnati, OH
45268.

-------
     4.3  A general yellow background color in the extract will limit the cell
          pathlength that can be employed and thus limit the sensitivity of
          the method.

5.   Apparatus and Materials

     5.1  Spectrophotometer-visible, scanning (510-370 nm).

     5.2  Separatory Funnels - 125 ml, 2000 ml,  250 ml.

     5.3  Cells - 1 to 5 cm pathlength, 20 ml volume maximum.

6.   Reagents, Solvents and Standards

     6.1  Ethyl acetate

     6.2  Hydrochloric acid -IN.  Add 83 ml cone, hydrochloric acid to
          water, and dilute to one liter.

     6.3  Chloramine T - 10% solution.  Prepare fresh daily by dissolving
          l.Og Chloramine T in 10 ml distilled water.

     6.4  Stock standard - 0.2 pg/yl.  Dissolve 100.0 mg purified benzidine
          in about 30 ml 1 N HC1.  Dilute to 500 ml with water.

7.   Preparation of Calibration Curve

     7.1  To a series of 125 ml separatory funnels, add 45 ml hydrochloric
          acid and 10 ml ethyl acetate.  Shake for one minute to saturate the
          acid layers.  Discard the solvent layers.  Dose the series with
          volumes from 1.0 to 20.0 Pi of stock standard, using syringes.

     7.2  Treat standards according to the Procedure beginning with 8.5.

8.   Procedure

     8.1  Adjust the sample pH to 8.5 - 9.0 with dilute NaOH or HC1.

     8.2  Transfer 1 liter of sample to a 2000 ml separatory funnel.  Add
          150 ml ethyl acetate and shake for two minutes.  Allow the layers to
          separate; then drain the water layer into a second 2-liter separatory
          funnel.  Drain the solvent layer into a 250 ml separatory funnel.

     8.3  Repeat the extraction of the water layer twice (i.e., total of
          three extractions) with 50-ml portions of ethyl acetate.   Combine
          all solvent layers; then discard the water layer.
                              J-

-------
     8.4  Extract the solvent layer three times with 15-ml portions of hydro-
          chloric acid by shaking 2 minutes and allowing the phases to separate.
          Combine the acid layers in a glass-stoppered container for cold
          storage until tine is available for analysis, or transfer the layers
          directly into a 125 ml separatory funnel.

     8.5  Prepare the spectrophotometer so that it is warmed and ready to use.
          The'remaining steps of the procedure must be performed rapidly on
          one sample at a time.

     8.6  To the hydrochloric acid solution in a 125 ml separatory funnel, add
          1.0 ml chloramine T solution, and mix.  Add 25.0 ml ethyl acetate
          with a pipet and shake for two minutes.  Allow the layers to separate;
          then discard the aqueous phase.

     8.7  Filter the solvent layer through coarse filter paper, and fill a 5
cm
          cell with the filtrate.

     8.8  Scan the solvent from 510 nra to 370 nir..  Use ethyl acetate for
          a blank with double beam instruments.  Use shorter pathlength cells
          in cases where absorbance exceeds 0.8.

9.   Calculation of Results

     9.1  Benzidine is identified by its absorbance maximum at 436 nm (figure
          1).  Dichlorobenzidine would appear to be similar, but has its
          absorbance maximum at 445 nm.

     9.2  Construct a baseline from the absorbance minimum at about 470 nm
          to the minimum at 390 nm (or 420 nm minimum for samples with a high
          background).  Record the absorbance of the peak maximum and the
          absorbance of the constructed baseline at the 436 nm.  Treat samples
          and standards in the same fashion.

     9.3  Using the net absorbance values, prepare a calibration plot from
          the standards.  Determine the total micrograms in each sample from
          this plot.

     9.4  Divide the  total micrograms by the sample volume, in liters, to
          determine ug/1.  Correct results for cell pathlength if necessary.

10.  Reporting Results

    10.1  Report results in micrograms per liter as benzidine, without correc-
          tion for recovery data.  When duplicate and spiked samples are
          analyzed, all data obtained should be  reported.

11.  Accuracy and Precision

    11.1  When 1-liter samples of Ohio River water were dosed with 1.80 up of
          benzidine,  an average of 1.24 ug was recovered.  The standard
          deviation was 0.092 yg/1  (n=8).

-------
12.  References
     (1)  Federal Register. Volume 39, Page 3779, Paragraph 1910.93; (January
          29, 1974).

     (2)  Classman, J. M., and Meigs, J. W., "Benzidine (A,4' - Diaminobiphenyl)
          and Substituted Benzidines", Arch. Industr. Hye-. 4. 519, (1951).

     (3)  Butt, L. T. and Strafford, N., "Papilloma of the Bladder in the
          Chemical Industry.  Analytical Methods for the Determination of
          Benzidine and B-Naphthylamine, Recommended by A. B. C. M. Sub-
          Comittee", J. Appl. Chem.. £., 525 (1956).

-------
Procedure Code:  3

Variations:  1-3

Title:  Method for Organochlorine Pesticides and Phthalate Esters in Industrial
        Effluents

1.   Summary

     This method is the EPA  (Federal Register, _3J3 No. 75, Pt. II) procedure
     for the pesticides and metabolites on the priority pollutants list.  The
     method uses liquid-liquid extraction to extract and concentrate the
     chlorinated pesticides and metabolites for gas chromatographic analysis.
     There is an optional Florisil cleanup step to separate phthalate esters
     (which are coextracted) from many of the pesticides.  Hence, the method
     can also be used to determine phthalate esters by selection of the proper
     Florisil fraction.

     The method variations in use are classified as follows:

     Variation      Effective
      Code No.        Date                   Description

         1          June 1977      Midwest" Research Method

         2          Feb. 1979      Southwest Research Method  (EPA Method #606)

         3          June 1977      Envirodyne Method, with GC conditions A-E

     Each of these variations is described in Section 6, Procedures.

2.   Interferences

     2.1  Solvents, reagents, glassware, and other sample-processing hardware
          must be demonstrated to be free from interfering compounds, which
          contribute gas chromatographic peaks or high baselines.   (See Refer-
          ence (1) for suggestions on reducing interferences).

     2.2  Care must be taken to ensure that the effect of a sample cleanup
          procedure upon the concentration of each analyte is known, and that
          the procedures for eliminating interferences with a cleanup step are
          applied consistently from sample to sample and on the  standards.

     2.3  Polychlorinated Biphenyls (PCB's) are a potential source of inter-
          ference in pesticide analysis.  The presence of PCB's  is  indicated
          by a large number  of partially-resolved or unresolved  peaks, which
          may occur throughout the entire chromatogram.   (See Procedure Code #9
          for  the determination of PCB's in the liquid-liquid extract from the
          pesticide-phthalate methods).

-------
     2.4  Organophosphorus pesticides that contain a nitro group (e.g.,
           parathion) also respond to the electron-capture detector, and nay
interfere with the determination of the organochlorine pesticides.  Such
compounds can be identified by their response to the flame photometric
detector (2).  Elemental sulfur and other organics with conjugated carbonyl
groups, nitro groups, or halogens will also interfere with the electron-
capture response to pesticides or phthalates.

3.   Apparatus and Materials

     3.1  Gas Chromatograph - Equipped with glass-lined injection port.

     3.2  Detector - For chlorinated pesticides, an electron-capture (EC)
          detector is required.  For phthalate esters, an EC or flame
          ionization detector  (FID) may be used.  The FID detector might be
          preferable to EC in the presence of a large quantity of EC -
          responsive interferences, where the phthalate concentrations are
          high enough to be measured accurately with FID.

     3.3  Recorder - Potentiometric strip chart (10-inch) compatible with the
          detector.

     3.4  Gas Chromatographic Column Materials - See Procedure section.

     3.5  Kuderna-Danish (K-D) Glassware (Kontes/

          3.5.1  Snyder Column - three ball  (macro) and two ball (micro)

          3.5.2  Evaporative Flasks - 500 ml

          3.5.3  Receiver ampules - 10 ml, graduated

          3.5.4  Ampule Stoppers

     3.6  Liquid Chromatography Column for Florisil - Chromaflex (400 mm long
          by 19 mm ID) with a  coarse fritted plate on the bottom and teflon
          stopcock; the column has a 250 ml reservoir bulb at the top of the
          column with flared-out funnel shape at the top of the bulb (Kontes
          K-420540-9011).

     3.7  Chromatographic Column - Pyrex (approximately 400 mm long by 20 mm
          ID) with coarse fritted plate on bottom.
     3.8  Blender - High-speed, glass or stainless steel cup.

          Florisil - PR Grade  (60/100 mesh); purchase activat
          store in the dark in glass containers with glass stoppers or foil-
          lined screw caps.  Before use, ac
          130 C in a foil-covered glass con
          value as described in Section 11.
3.9  Florisil - PR Grade (60/100 mesh); purchase activated at 1250°F and
     store in the dark in glass containers with glass stoppers or foj
     lined screw caps.  Before use, activate each batch overnight at
     130 C in a foil-covered glass container; determine the lauric acid

-------
4.   Reagents, Solvents and Standards

     4.1  Ferrous Sulfate - (ACS) 30% solution in distilled water.

     4.2  Potassium Iodide - (ACS) 10% solution in distilled water.

     4.3  Sodium Chloride - (ACS) Saturated solution in distilled water  (pre-
          rinse NaCl with hexane).

     4.4  Sodium Hydroxide - (ACS) 10 N in distilled water.

     4.5  Sodium Sulfate - (ACS) Granular, anhydrous (conditioned at 400°C for
          4 hours).

     4.6  Sulfuric Acid - (ACS) Mix equal volumes of concentrated H.SO, with
          distilled water.                                         i  *

     4.7  Diethyl Ether - Nanograde, redistilled in glass, if necessary.

          4.7.1  Must contain 2 percent alcohol and be free of peroxides by
                 following test:  To 10 ml of ether in glass-stoppered cylinder,
                 previously rinsed with ether, add 1 ml of freshly-prepared
                 10 percent RI solution.  Shake and let stand 1 minute.  No
                 yellow color should be observed in either layer.

          4.7.2  Decompose ether peroxides by adding 40 g of 30 percent ferrous
                 sulfate solution to each liter of solvent.  CAUTION;  Reaction
                 may be vigorous if the solvent contains a high concentration
                 of peroxides.

          4.7.3  Distill deperoxidized ether in glass and add 2 percent ethanol.

     4.8  Hexane, Methylene Chloride, Petroleum Ether (boiling range 30  to
          60 C) - Nanograde,  redistill in glass if necessary.

     4.9  Phthalate or Pesticide - Compound being analyzed (99+Z purity).

5.   Calibration and Quality Control

     5.1  Gas chromatographic operating conditions are considered acceptable if
          the response to dicapthon is at least 50 percent of full scale when
          0.06 ng (or less) is injected for electron-capture detection.  For
          all quantitative measurements, the detector must be operated within
          its linear response range, and the detector noise level should be
          less than 2 percent of full scale.

     5.2  Standards are injected frequently as a check on the stability of
          operating conditions.

     5.3  Duplicate and spiked sample analyses are used for quality control
          checks.  Each time a set of samples is extracted, a method blank is
          determined on a volume of distilled water equal to that used to
          dilute the sample.

-------
6.   Procedures

     6.1  Variation #1 (Midwest Research Phthalate Method):

          Adjust the pH of a 1 liter sample to 7.0.  Place the sample in a
          heavier-than-water liquid-liquid extractor, and extract with methylene
          chloride overnight (16-20 hr.).  Pass the extract through a Na.SO,
          drying column and evaporate to approximately 10 ml in a K-D macro
          apparatus on a steam bath.  Evaporate the samples and exchange with
          hexane, to a small volume under a stream of dry nitrogen.  Then
          clean up the sample on a Florisil column according to Section 7, but
          discard the 6 percent ether in petroleum ether fraction and elute
          the column with a 50:50 (v:v) mixture of ethyl ether and petroleum
          ether for phthalate recovery.  The gas chromatographic conditions
          are:

               GC Column:     1% SP-2250 on 100/120 mesh Supelcoport in a
                              6 ft. x A mm ID (glass) column.
               Temperature:   215°C Isothermal    ,,
               Detector:      Electron capture, Ni   at 300°C.
               Carrier:       Argon/methane at 60 ml/min.

     6.2  Variation #2 (Southwest Research Phthalate Method; same as E?A
          Method #606 (3)).

          Measure 1-liter of sample into a graduated cylinder.  Pour the
          sample into a 2-liter separatory funnel and adjust the pH to 11 or
          above with 6N NaOH, using pH paper indicator.  Extract 3 times with
          60 ml of methylene chloride, and combine the extracts in a 500 ml K-
          D evaporator.  Evaporate to about 5 ml on a steam bath, add 100 ml
          of hexane, and evaporate to about 2 ml, using a nitrogen gas stream
          for the final evaporation.  To clean up the extract for phthalates,
          take the following steps:

          Prepare Florisil by heating to 400°C overnight.  Then add 3% HJ)
          (m/m) and mix thoroughly.  Let stand for at least 2 hours.  Use 10
          mm ID column chromatography tube with glass frit and teflon stopcock.
          (Kontes Catalog K420540-0213).

          1.   Add 10 g of Florisil (3% H20) to the 1 cm ID chromatography
               tube and tap to settle.

          2.   Add   1 cm Na2SO, to top of column.

          3.   Add hexane to tube and pre-elute 40 ml.  Discard.

          4.   Drain until solvent level is even with top of Na.SO,.
                                                               24
          5.   Add sample in   2 ml hexane to top of column.

          6.   Drain sample into column (to top of Na.SO,).
                                  T-  ii

-------
     7.   Wash sides of tube with = 2 ml hexane and drain to top of Na SO,.

     8.   Add hexane to top of column and eiute 60 ml of hexane.
          Discard this.

     9.   Drain hexane to top of Na^SO,  and add * 2 ml of 20% ether/hexane.

    10.   Drain new solvent to top of Na2SO,.  Add more 20% ether hexane.

    11.   Place a 500 ml Kuderna-Danish evaporator below column.

    12.   Collect 100 ml of the 20% ether/hexane eluate in the K-D evaporator.

    13.   Evaporate the above eluate to  =10 ml on a steam bath.

    14.   Transfer to a solvent rinsed 4-dram vial with teflon cap liner.

    15.   Adjust final volume of concentrate to 5 ml.  Evaporate under
          nitrogen if necessary.

    16.   For GC analysis a 1:20 to 1:100 dilution of this fraction may be
          necessary.

     The gas chromatographic conditions are:

          GC Column:     6' x 1/4" 1.5% SP-2250/1.95% SP-2401 on 100/120
                         mesh Supelcoport.
          Temperature:   180°C for 16 min,  then heat rapidly to 220°C and
                         hold for rest of run (20 min).
          Detector:      Electron capture,  300 C.
          Carrier:       Argon/methane/60 cc/min

6.3  Variation #3 (Envirodyne Method for Pesticides and Phthalates):

     Blend the sample if suspended matter is present, and adjust the pH to
     6.5 to 7.5 with 50 percent sulfuric acid or 10 N sodium hydroxide.
     Extract 1 liter of sample three times with 60 ml portions of 15
     percent methylene chloride in hexane (v:v) in a 2-liter separatory
     funnel, shaking for 2 min. each time.   Collect the solvent layers
     from the three extractions in a 500 ml K-D flask after drying the
     extracts by passage through a column (3-4 inches) of anhydrous sodium
     sulfate.  Concentrate the extracts to 6-10 ml into the (10 ml) K-D
     ampule, over a steam bath.

     If Florisil cleanup is necessary for determining pesticides/phthalates
     without excessive interference from each other or other compounds,
     use the (EPA) Florisil procedure in Section 7.

-------
Gas chromatography is performed as follows:
A)   GC Column:

     Temperature:
     Inj ector:
     Detector:
     Carrier:
                         3% SP-1000 on 100/120 mesh Supelcoport in
                         a 6* x 2 mm ID glass column.
                         100° to 250°C at 8°/min.
Ni
                              at 325°C
                         Argon/methane at 60 ml/min.
          Note:  (Specified for phthalate esters)
Other columns which have been used for phthalates or pesticides are:
     B)   GC Column:

          Temperature:
          Injector:
          Detector:
          Carrier:

     C)   GC Column:

          Temperature:
          Injector:
          Detector:
          Carrier:

     D)   GC Column:

          Temperature:
          Injector:
          Detector:
          Carrier:
3% QF-1 on Chromosorb WHP (00/100 mesh) in
a 6' x 2 mm ID glass column.
215°C Isothermal
                    Ni
     at 325°C
Argon/methane at 60 ml/min.

3% SP-1000 on 100/120 mesh Supelcoport in
a 6* x 2 mm ID glass column.
220°C Isothermal
     at 270C
Argon/methane at 60 ml/min.

3% SP-1000 on 100/120 mesh Supelcoport in a
6* x 2 mm ID glass column.
100°C to 250°C at 8 C/min.
Ni   at 325C
Argon/methane at 60 ml/min.
E)   GC Column:

     Temperature:
     Injector:
     Detector:
     Carrier:

F)   GC Column:

     Temperature:
     Injector:
     Detector:
     Carrier:
                         7% OV-17 + 9% OV-210 on Chromosorb W in a
                         6* x 2 mm ID glass column.
                         190°C Isothermal
                         Ni"" at 325"C
                         Argon/methane at 60 ml/min.

                         1.5% SP-2250 -I- 1.95% SP-2401 on 100/120
mesh Supelcoport in a 6*
210°C Isothermal
                                                  x 2 mm glass column.
                         Ni   at 325C
                         Argon/methane at 60 ml/min.
          (Date:  May 1979)
     G)   GC Column:

          Temperature:
          Injector:
          Detctor:
          Carrier:
1.5% SP-2250 + 1.95% SP-2401 on 100/120 mesh
Supelcoport in a 6' x 2 mm glass column.
190°C Isothermal
                    Ni
     at 325°C
          (Date:  May 1979)
Argon/Methane at 40 ml/min.

-------
7.   Florisil Column Adsorption Chromatography

     7.1  Adjust the sample extract volume to 10 ml.

     7.2  Place a charge of activated Florisil (weight determined by lauric-
          acid value (see Section 10) in a Chromaflex column.  After settling
          the Florisil by tapping the column, add about one-half inch layer of
          anhydrous granular sodium sulfate to the top.

     7.3  Fre-elute the column, after cooling, with 50 to 60 ml of petroleum
          ether.  Discard the eluate and just prior to exposure of the sulfate
          layer to air, quantitatively transfer the sample extract into the
          column by decantation and subsequent petroleum ether washings.
          Adjust the elution rate to about 5 ml per minute and, separately,
          collect up to three eluates in 500 ml K-D flasks equipped with 10 ml
          ampules.  Perform the first elution with 200 ml of 6 volume percent
          of peroxide-free ethyl ether (containing 2 percent ethanol) in
          petroleum ether.  Perform the second elution with 200 ml of 15
          volume percent of peroxide-free ethyl ether  (containing 2 percent
          ethanol) in petroleum ether.  Elute a third tine with 200 ml of 50
          volume percent of peroxide-free ethyl ether  (containing 2 percent
          ethanol) in petroleum ether.

     7.4  Eluate composition - By using an equivalent quantity of any batch of
          Florisil as determined by Its lauric acid value, pesticides, phthalates
          and PCB's will be separated into the eluates indicated in the following
          lists.  However, the distribution of the compounds of Interest
          should be checked by performing the florisil clean-up with a mixed
          standard of these compounds at a concentration which is readily
          quantified on the GC.

                    62 Eluate

                    Aldrin         Heptachlor               Pentachloro-
                    BHC            Heptachlor expoxide       nitrobenzene
                    Chlordane      Lindane                  Strobane
                    DDD            Methoxychlor             Toxaphene
                    DDE            Mirex                    Trifluralin
                    DDT                                     PCB's

                    15Z Eluate               50% Eluate

                    Endosulfan I             Endosulfan II
                    Endrin                   Captan
                    Dieldrin
                    Dichloran
                    Phthalate  esters

     7.5  Concentrate  the extracts to approximately 5  ml  in  the K-D evaporator
          in a  hot water bath  and add a small amount of sodium sulfate  to
          assure  a water-free  sample.
                             T-/1

-------
     7.6  Analyze by gas chromatography.
                                 •

8.   Calculation of Results

     8.1  Determine the phthalate ester or pesticide concentration by using
          the following absolute calibration procedure or the relative cali-
          bration procedure described in Part 1, Section 3.4.2 of Reference (1).

                                       (A)  (B)  (V )
                    Micrograms/liter »
                                         
-------
10.3  Reagents and Solvents

      10.3.1   Alcohol, ethyl —USP or absolute, neutralized to phenol-
               phthalein,
      10.3.2   Hexane — Distilled from all glass apparatus.
      10.3.3   Laurie acid -- Purified, CP.
      10.3.4   Laurie acid solution — Transfer 10.000 g lauric acid to a
               500 ml volumetric flask, dissolve in hexane, and dilute to
               500 ml (1 ml = 20 mg).

      10.3.5   Phenolphthalein Indicator — Dissolve 1 g in alcohol and
               dilute to 100 ml.

      10.3.6   Sodium hydroxide — Dissolve 20 g NaOH (pellets, reagent
               grade) in water and dilute to 500 ml (1 BI).  Dilute 25 ml
               1 N NaOH to 500 ml with water (0.05 N).  Standardize as
               follows:
               Weigh 100-200 mg lauric acid into 125 ml Erlenmeyer flask.
               Add 50 ml neutralized ethyl alcohol and 3 drops phenol-
               phthalein indicator; titrate to permanent end point.  Cal-
               culate mg lauric acid/ml 0.05 II NaOH (about 10 mg/ml).

10.4  Procedure

      10.4.1   Transfer 2.000 g Florisil to 25 ml glass stoppered Erlen-
               meyer flasks.  Cover loosley with aluminum foil and heat
               overnight at 130 C.  Stopper, cool to room temperature,
               add 20.0 ml lauric acid solution (400 mg), stopper, and
               shake occasionally for 15 min.  Let adsorbent settle and
               pipet 10.0 ml of supernatant into 125 ml Erlenmeyer flask.
               Avoid inclusion of any Florisil.

      10.4.2   Add 50 ml neutral alcohol and 3 drops indicator solution;
               titrate with 0.05 1$ NaOH to a permanent end point.

10.5  Calculation of Lauric Acid Value and Adjustment of Column Weight

      10.5.1   Calculate amount of lauric acid adsorbed on Florisil as
               follows:
               Lauric Acid value » mg lauric acid/g Florisil » 200 -  (ml
               required for titration X mg lauric acid/ml 0.05 N NaOH).

      10.5.2   To obtain an equivalent quantity of any batch of Florisil,
               divide 110 by lauric acid value for that batch and multiply
               by 20 g.  Verify proper elution of pesticides and phthalates
               by 13.6.

10.6  Test  for Proper Elution Pattern and Recovery of Pesticides and
      phthalates.  Prepare a test mixture containing aldrin, heptachlor
      epoxide, p,p' -DDE, phthalates, dieldrin, parathion and malathion.
      Dieldrin and parathion and phthalates should eluta in the 15% eluate;
      all but a  trace of malathion  in  the 50% eluate and the others in
      the 6% eluate.

-------
11.  References

     1.   "Method for Organic Pesticides in Water and Wastewater", Environmental
          Protection Agency, National Environmental Research Center, Cincinnati,
          Ohio (1971).

     2.   "Method for Organopnosphorus Pesticides in Industrial Effluents",
          Environmental Protection Agency,  National Environmental Research
          Center, Cincinnati, Ohio (1973).

     3.   Methods for Organic Compounds in Municipal and Industrial Wastewater.
          U.S. EPA Environmental Monitoring and Support Laboratory, Cincinnati,
          Ohio, March, 1979.

     4.   "Pesticide Analytical Manual", U.S. Dept. of Health, Education and
          Welfare, Food and Drug Administration, Washington, D.C., Volumes I
          and II (1968).

     5.   Mills, P.A., "Variation of, Florisil Activity:  Simple Method for
          Measuring Adsorbent Capacity and its Use in Standardizing Florisil
          Columns", Journal A.O.A.C.. 51, 29 (1968).

-------
Procedure Code:  4

Variations:  1-2

Title:    Total Cyanide

1.   Summary

     The determination of simple cyanides plus complex cyanides in wastewaters
may be performed by the EPA (1) procedure, which uses an acid distillation
with cuprous chloride, or by the procedure in Standard Methods (2), which uses
an acid distillation with magnesium chloride and mercuric chloride. Recent
studies with coke plant effluents (3) showed a) that the use of cuprous
chloride gave lower results than the use of magnesium chloride, and b) the
presence of high thiocyanate in the samples led to sulfide in the cyanide
absorbing trap.  This sulfide was removed by precipitation with cadmium
carbonate, followed by filtration, before final cyanide analysis.

     In order to follow the method variations lor cyanide, the following table
should be used:

     Variation      Effective
      Code No.        date          Method Source

       1            June 1977       EPA (1)

       2            June 1977       Standard Methods (2)

     With either method, the cyanide can be determined by titration with
silver nitrate or colorimetry with pyridine-pyrazolone or pyridine-barbituric
acid reagents, as described in the methods referenced.

2.   References
     1.   Methods for Chemical Analysis of Water and Wastes, U.S. EPA Office
          of Technology Transfer, Washington, D.C. (1974).

     2.   Standard Methods for the Examination of Water and Wastewater,
          14th Edition, American Public Health Assn., Washington, D.C. (1976).

     3.   Barton, P. J., Hammer, C.A., and Kennedy, D.C., "Analysis of Cyanides
          in Coke Plant Wastewater Effluents, "JWPCF, 50, 234-239  (1978).

-------
Procedure Code:  5

Variations:  1 to 3

Title:  A-26 Resin/GC-FID Method for Phenols

1.   Summary

          The method given here is that reported by Chriswell e_t al.  (1).
     phenols are trapped (as phenolate anions) on A-26 anion exchange  resin by
     passing alkaline sample through a column of resin in the hydroxyl form.
     Any neutral organic compounds retained by the resin are washed off with
     alkaline methanol.  The column is then washed with aqueous hydrochloric
     acid to convert the phenolate anions back to the acid fora and elute them
     from the column.  The elution is completed with an acetone-water  solution
     and all eluates extracted with methylene chloride for gas chromatography
     with flame ionization detection (FID).

2.   Interferences

     2.1  Solvents, reagents, glassware, and other sample-processing hardware
          may yield discrete artifacts and/or elevated baselines that  could
          cause misinterpretation of gas chromatograms.  All of these materials
          must be demonstrated to be free from interferences under the conditions
          of the analysis   Specific selection of reagents and purification of
          solvents by distillation in all-glass systems may be required.

     2.2  Common inorganic ions and many organic substances cause no inter-
          ference, and neutral organics that are retained by the resin can be
          removed by a methanol wash.

     2.3  The interferences in industrial effluents are many and varied,
          and often pose great difficulty in obtaining accurate and precise
          measurement of phenols.  Sample cleanup procedures are generally
          required and may result in the loss of certain phenols.  Therefore,
          great care should be exercised in the selection and use of methods
          for eliminating or minimizing interferences.  It is not possible to
          describe procedures for overcoming all the interferences that may be
          encountered in industrial effluents.

3.   Apparatus and Materials

     3.1  Gas Chromatograph - equipped with dual flame ionization detectors.

     3.2  Recorder - Potentiometric strip chart (10 inch) compatible with
          the detectors.

     3.3  Gas Chromatographic Column Materials - see method variations

-------
     3.4  Detector Options

          3.A.I  Dual flame ionization

          3.4.2  Electron capture - radioactive (tritium or nickel 63) - if
                 phenols of interest are halogenated.

     3.5  Kuderna-Danish (K-D) Glassware

          3.5.1  Snyder column - three ball (macro) and two ball  (micro)

          3.5.2  Evaporative flasks - 500 ml

          3.5.3  Receiver ampules - 10 ml, graduated

          3.5.4  Ampule stoppers

     3.6  Chromatographic column - 1/2 in x 6 in glass chromatographic column
          with A 1,000 ml reservoir and a coarse fritted plate on bottom,
          prepared by pouring a water slurry of the cleaned A-26 resin into the
          column until the resin bed is within about one-half inch of the top
          of the column.  The column is placed in the hydroxy form by passing
          approximately 20 ml of 0.1 M sodium hydroxide solution through the
          resin.  Excess sodium hydroxide is washed from the resin bed with 50
          ml of purified distilled water.  Verify that there was excess hydrox-
          ide on the column by testing the first portion of water wash with pH
          paper (color should indicate pH^9).

     3.7  Filter - medium porosity, 150 ml sintered glass filter

     3.8  Separatory funnels - 125 ml with Teflon stopcock

     3.9  Graduated cylinders - 50 ml, 100 ml and 500 ml

     3.10  Porous polymer resin - Rohm & Haas A-26 anion exchange resin screened
           to remove beads smaller than 60 mesh.  Clean new resin by placing
           in a sintered glass filter attached to a suction flask, washing
           successively with 2 M sodium hydroxide, purified distilled water,
           4 M hydrochloric acid, purified distilled water and acetone.
           Repeat sequence of washings until no color is apparent in the final
           acetone wash (generally one to three times).  Then extract with
           acetone for 24 hours in a Soxhlet extractor.

4.   Reagents, Solvents and Standards

     4.1  Sodium hydroxide - (ACS)

          4.1.1  2 M solution in distilled water

          4.1.2  0.1 M solution in distilled water

-------
     4.2  Hydrochloric acid - (ACS) 4 M solution in distilled water

     4.3  Acetone, methylene chloride, methanol-nanograde, redistilled in
          glass if necessary.

     4.4  Hydroxylamine hydrochloride - ACS

     4.5  Sodium hydrosulfite - ACS

     4.6  Phenol - compound(s) being analyzed

5.    Quality Control

     5.1  Standards should be injected frequently, as a check on the stability
          of the operating conditions.

     5.2  Duplicate and spiked sample analyses are recommended as quality
          control checks.   When the routine occurrence of a phenol is being
          observed, the use of quality control charts is recommended.

     5.3  Each time a set  of samples is extracted, a method blank is determined
          on .a volume of distilled water equivalent to that used to dilute the
          sample.

6.    Sample Preparation

     6.1  Take a 500 ml sample of water.  If phenol standards are added to
          water containing chlorine, add 15 to 25 mg of hydroxylamine hydro-
          chloride and allow to stir for at least 5 minutes before proceeding.
          Add 15 to 25 mg of sodium hydrosulfite to the sample, and adjust the
          pH to between 12.0 and 12.5 with 2 M sodium hydroxide.

     6.2  If a precipitate forms, coagulate it by allowing the sample to sit
          for about 15 minutes.  Decant the supernatant liquid through a
          medium-porosity, 150 ml sintered-glass filter attached to a suction
          flask.  Wash the precipitate into the filter with a minimum amount
          of water, then wash thoroughly with approximately 50 ml of distilled
          water.

     6.3  Pour the filtered water and washings into the reservoir of the
          adsorption column and allow to flow through the resin column at a
          rate of 10 to 15 ml/min.*  When the liquid level reaches the top of
          the resin bed, wash the column with 25 ml of basic methanol (2 ml of
          2 M sodium hydroxide in 23 ml of methanol') and 25 ml of distilled
          water.

     6.4  Place a 125-ml separatory funnel under the column, and elute the
          column with 25 ml of 4 M hydrochloric acid, then with 25 ml of
          distilled water.  (The first portion of water should test acidic to
          pH paper).  Extract the solution in the separatory funnel with 25 ml
          of methylene chloride.  Allow the phases to separate and become clear;
          then drain the lower methylene chloride layer into a second 125-ml
          separatory funnel and discard the aqueous layer.

*Note:  Some workers have found better phenolate exchange when a slower flow
        rate is used (about 3 ml/min.) in this step (see Variation #2).

-------
     6.5  Elute the column with 60 ml of 5:1 (V:V) acetone'.water and 50 ml of
          distilled water in sequence into the separatory funnel containing
          the methylene chloride.  Shake, allow the phases to separate and
          clear; then drain the lower organic layer into a K-D evaporating
          flask.

     6.6  Add a small boiling chip to the flask, attach a bnyder column, and
          evaporate the solvent over a steam bath until the volume is reduced
          to approximately 0.5 ml.  Remove the evaporating flask from the
          steam, and immediately spray the outside with acetone to condense the
          vapors inside.   Adjust the volume to 'exactly 1.0 ml with acetone.

     6.7  Inject 2 yl of  the acetone solution into the gas chromatograph (see
          GC conditions,  below).

          Note:     If the acetone solvent peak interferes with some of the
                    individual phenol peaks, the acetone may be removed by
                    adding approximately 7 ml of pentane to the evaporated (1.0
                    ml) acetone extract.  By evaporating and diluting to exactly
                    1.0 ml with pentane, the acetone is volatilized by azeotropic
                    distillation.  Injection of a pentane solution permits
                    operation of the gas chromatograph at significantly lower
                    attenuation settings without interference from a tailing
                    edge  of the solvent peak.

7.   Calculation of Results

     7.1  Determine the phenol concentration by comparing either peak heights
          or areas against a previously-prepared calibration curve.

8.   Reporting Results

     8.1  Report results  in micrograms per liter without correction for recovery
          data.  When duplicate and spiked samples are analyzed, all data
          obtained must be reported on the analytical quality control report
          sheets.

9.   Variations to Procedure Code #5
     Variation
      Code No.
Effective
  Date

June 1977
Description

GC Column:



Temperature:


Injector:

Detector:
                                                    1% SP-1240 DA on 100/120
                                                    mesh Supelcoport in a 4 ft
                                                    x 2 mm ID glass column.

                                                       80°C to 180°C at 10°/min.,
                                                       hold at 180°C for 15 min.
                                                    240°C
                                 I-.23-
                                                    FID at 240 C

                                        Carrier:    He at 35 ml/min.

                                        (Source:  Envirodyne Engineers, Inc.)

-------
     Variation
      Code No.
Effective
  Date

February 1979
                    June 1977
Description

Same as Variation 1, except Column
Temperature:  80  for 5 min.,
                                                      80°  to 180° at  10°/min.t
                                                      hold at 180° for  15 min.
                    3 ml/min flow rate for initial phenolate
                    anion exchange (Step 6.3); regenerate
                    resin with 2M NaOH instead of 0.1M
                    NaOH.
                                        GC column:  1% SP-1240 DA on Supelcoport
                                        (100/120 mesh) in a 6' x 2mm ID glass
                                        column.

                                        Temperature:  4 min. at 85 C, then
                                                      10°/min to 200 C

                                        Injector:  180°C

                                        Detector:  FID at 200°C

                                        Carrier:  N- at 30 ml/min

                                        (Source:  Midwest Research Institute)
10.  References

     1.   Chriswell, C. D., Chang, R. C., and Fritz, J. S., "Chronatographic
          Determination of Phenols in Water", Anal. Chem.. 47, 1325-1329
          (1975).

-------
                                                                      ^ In I 71
Procedure Code:  5

Variations:  1 to 5

Title:  A-26 Resin/GC-FID Method for Phenols

1.   Summary

          The method given here is that reported by Chrisvell et, al. .(1) •
     Phenols are trapped (as phenolate anions) on A-26 anion exchange resin by
     passing alkaline sample through a column of resin in the hydroxyl form.
     Any neutral organic compounds retained by the resin are washed off with
     alkaline methanol.  The column is then washed with aqueous hydrochloric
     acid to convert the phenolate anions back to the acid form and elute them
     from the column.  The elution is completed with an acetone-water solution
     and all eluates extracted with methylene chloride for gas chromatography
     with flame ionization detection (FID).

          For other methods used for phenol analysis, see Procedure Codes #7
     and 915.

2.   Interferences

     2.1  Solvents, reagents, glassware, and other sample-processing hardware
          may yield discrete artifacts and/or elevated baselines that could
          cause misinterpretation of gas chromatograms.  All of these materials
          must be demonstrated to be free from interferences under the conditions
          of the analysis.  Specific selection of reagents and purification of
          solvents by distillation in all-glass systems may be required.

     2.2  Common inorganic ions and many organic substances cause no inter-
          ference, and neutral organics that are retained by the resin can be
          removed by a methanol wash.

     2.3  The interferences in industrial effluents are many and varied,
          and often pose great difficulty in obtaining accurate and precise
          measurement of phenols.  Sample cleanup procedures are generally
          required and may result in the loss of certain phenols.  Therefore,
          great care should be exercised in the selection and use of methods
          for eliminating or minimizing interferences.  It is not possible to
          describe procedures for overcoming all the interferences that may be
          encountered in industrial effluents.

3.   Apparatus and Materials

     3.1  Gas Chromatograph - equipped with dual flame ionization detectors.

     3.2  Recorder - Potentiometric strip chart  (10 inch) compatible with
          the detectors.

     3.3  Gas Chromatographic Column Materials - see method variations
                                >2 A

-------
     3.4  Detector Options

          3.4.1  Dual flame ionization

          3.4.2  Electron capture - radioactive  (tritium or nickel 63) - if
                 phenols of interest are halogenated.

     3.5  Kuderna-Danish (K-D) Glassware

          3.5.1  Snyder column - three ball (macro) and tvo ball  (micro)

          3.5.2  Evaporative flasks - 500 ml

          3.5.3  Receiver ampules - 10 ml, graduated

          3.5.4  Ampule stoppers

     3.6  Chromatographic column - 1/2 in x 6 in glass chromatographic column
          with a 1,000 ml reservoir and a coarse fritted plate on bottom,
          prepared by pouring a water slurry of the cleaned A-26 resin into the
          column until the resin bed is within about one-half inch of the top
          of the column.  The column is placed in the hydroxy form by passing
          approximately 20 ml of 0.1 M sodium hydroxide solution through the
          resin.  Excess sodium hydroxide is washed from the resin bed with 50
          ml of purified distilled water.  Verify that there was excess hydrox-
          ide on the column by testing the first portion of water wash with pH
          paper (color should indicate pHj>9).

     3.7  Filter - medium porosity, 150 ml sintered glass filter

     3.8  Separatory funnels - 125 ml with Teflon stopcock

     3.9  Graduated cylinders - 50 ml, 100 ml and 500 ml

     3.10  Porous polymer resin - Rohm & Haas A-26 anion exchange resin screened
           to remove beads smaller than 60 mesh.  Clean new resin by placing
           in a sintered glass filter attached to a suction flask, washing
           successively with 2 M sodium hydroxide, purified distilled water,
           4 M hydrochloric acid, purified distilled water and acetone.
           Repeat sequence of washings until no color is apparent in the final
           acetone wash (generally one to three times).  Then extract with
           acetone for 24 hours in a Soxhlet extractor.

4.   Reagents, Solvents and Standards

     4.1  Sodium hydroxide - (ACS)

          4.1.1  2 M solution in distilled water

          4.1.2  0.1 M solution in distilled water

-------
     4.2  Hydrochloric acid - (ACS) 4 M solution in distilled water

     4.3  Acetone, methylene chloride, methanol-nanograde, redistilled in
          glass if necessary.

     4.4  Hydroxylamine hydrochloride - ACS

     4.5  Sodium hydrosulfite - ACS

     4.6  Phenol - compound(s) being analyzed

5.   Quality Control

     5.1  Standards should be injected frequently, as a check on the stability
          of the operating conditions.

     5.2  Duplicate and spiked sample analyses are recommended as quality
          control checks.  When the routine occurrence of a phenol is being
          observed, the use of quality control charts is recommended.

     5.3  Each time a set of samples is extracted, a method blank is determined
          on a volume of distilled water equivalent to that used to dilute the
          sample.

6.   Sample Preparation

     6.1  Take a 500 ml sample of water.  If phenol standards are added to
          water containing chlorine, add 15 to 25 mg of hydroxylamine hydro-
          chloride and allow to stir for at least 5 minutes before proceeding.
          Add 15 to 25 mg of sodium hydrosulfite to the sample, and adjust the
          pH to between 12.0 and 12.5 with 2 M sodium hydroxide.

     6.2  If a precipitate forms, coagulate it by allowing the sample to sit
          for about 15 minutes.  Decant the supernatant liquid through a
          medium-porosity, 150 ml sintered-glass filter attached to a suction
          flask.  Wash the precipitate into the filter with a minimum amount
          of water, then wash thoroughly with approximately 50 ml of distilled
          water.

     6.3  Pour the filtered water and washings into the reservoir of the
          adsorption column and allow to flow through the resin column at a
          rate of 10 to 15 ml/min.*  When the liquid level reaches the top of
          the resin bed, wash the column with 25 ml of basic methanol (2 ml of
          2 M sodium hydroxide in 23 ml of methanol) and 25 ml of distilled
          water.

     6.4  Place a 125-ral separatory funnel under the column, and elute the
          column with 25 ml of 4 M hydrochloric acid, then with 25 ml of
          distilled water.   (The first portion of water should test acidic to
          pH paper).  Extract the solution in the separatory funnel with 25 ml
          of methylene chloride.  Allow the phases to separate and become clear;
          then drain the lower methylene chloride layer into a second 125-ml
          separatory funnel and discard the aqueous layer.

*Note:  Some workers have found better phenolate exchange when a slower flow
        rate is used (about 3 ml/min.) in this step (see Variation #2).

-------
     6.5  Elute the column with 60 ml of 5:1 (V:V) acetonerwater and 50 ml of
          distilled water in sequence into the separatory funnel containing
          the methylene chloride.  Shake, allow the phases to separate and
          clear; then drain the lower organic layer into a K-D evaporating
          flask.

     6.6  Add a small boiling chip to the flask, attach a Snyder column, and
          evaporate the solvent over a steam bath until the volume is reduced
          to approximately 0.5 ml.  Remove the evaporating flask from the
          steam, and immediately spray the outside with acetone to condense the
          vapors inside.  Adjust the volume to exactly 1.0 ml with acetone.

     6.7  Inject 2 yl of the acetone solution into the gas chromatograph (see
          GC conditions, below).

          Note:     If the acetone solvent peak interferes with some of the
                    individual phenol peaks, the acetone may be removed by
                    adding approximately 7 ml of pentane to the evaporated (1.0
                    ml) acetone extract.  By evaporating and diluting to exactly
                    1.0 ml with pentane, the acetone is volatilized by azeotropic
                    distillation.  Injection of a pentane solution permits
                    operation of the gas chromatograph at significantly lower
                    attenuation settings without interference from a tailing
                    edge of the solvent peak.

7.   Calculation of Results

     7.1  Determine the phenol concentration by comparing either peak heights
          or areas against a previously-prepared calibration curve.

8.   Reporting Results

     8.1  Report results in micrograms per liter without correction for recovery
          data.  When duplicate and spiked samples are analyzed, all data
          obtained must be reported on the analytical quality control report
          sheets.

9.   Variations to Procedure Code #5

     Variation      Effective
      Code No.        Date              Description

         1          June 1977           GC Column:  1% SP-1240 DA on 100/120
                                                    mesh Supelcoport in a 4 ft
                                                    x 2 mm ID glass  column.

                                        Temperature:   80°C to 180°C at 10°/min.,
                                                       hold at 180°C for 15 min.

                                        Injector:    240°C

                                        Detector:    FID at 240°C

                                        Carrier:    He at 35 ml/min.

                                        (Source:  Envirodyne Engineers,  Inc.)

                              TOT-

-------
    mVariation
      Code No.
Effective
  Date

February 1979
                    June 1977
A.
     Description
                                        Same as Variation 1,  except Column
                                        Temperature:   80  for 5 min.,
                                                      80° to  180° at 10°/min.,
                                                      hold at 180° for 15 min.
                    3 ml/min flow rate for initial phenolate
                    anion exchange (Step 6.3); regenerate
                    resin with 2M NaOH instead of 0.1M
                    NaOH.
                    May 1979
                     May 1979
               B.   GC column:  1% SP-1240 DA on Supelcoport
                    (100/120 mesh) in a 6' x 2mm ID glass
                    column.

                    Temperature:  4 min. at 85°C. then
                                  10°/min to 200- C

                    Injector:  180°C

                    Detector:  FID at 200°C

                    Carrier:  N- at 30 ml/min

                    (Source:  Midwest Research Institute)

                    Same as Variation 1, except as follows:
                    Column Temperature:  80°C for 10 min.,
                                         increase to 180
                                         at 5 /min., hold
                                         at 180° for 30
                                         minutes

                    Injector:  190°C

                    Detector:  FID at 300°C

                    Carrier:  He at 35 ml/min.

                    (Source:  EEI)

                    Same as Variation 1, except as follows:
                    Column Temperature:  80°C for 13 min.,
                                         increase to 180°C
                                         at 10°/min.. and
                                         hold at 180 C for
                                         30 min.
10.  References

     1.   Chriswell, C. D., Chang, R. C., and Fritz, J. S., "Chromatographic
          Determination of Phenols in Water", Anal. Chem.. 47. 1325-1329
          (1975).

-------
                        FV      ':r*rr^
Procedure Code:   7      i V i       tv            5/'~>/79

Variations:  1-6

Title:    Microextraction Method  for Organic Compounds
          in Industrial Effluents
          (Source:  Southwest Research Institute)

1.   Summary

     The method  given here is adapted from that of Henderson et al.  (!)• (See
     Procedure Code #14).   The method is similar to that of Procedure Code #14,
     except that the solvent/sample ratio is much smaller with the microextraction
     method than with Procedure //14.  In both methods a water sample is equili-
     brated once with an organic  solvent.  In microextraction the water to
     solvent ratio may be as  high as 1000:1, but generally is in the range of
     200:1 to 10:1.  The water sample may or may not be saturated with salt as
     required.

     The method  variations are classified as follows:

     Variation           Effective
      Code No.             Date              Description

         1               Feb.  1979     Benzene-Toluene-Ethyl Benzene procedure

         2              March 1979    Nitrobenzene procedure

         3              May  1979      Behzene-Toluerie-Ethyl Benzene procedure

         4              May  1979      Phthalates procedure

         5               May  1979      Phenols procedure
                         May 1979       Polynuclear Aromatic Hydrocarbon
                                       procedure
2.   Sample Collection
     Samples to be analyzed for volatile organic compounds are collected in 40
     ml serum vials.   The vials are  filled to overflowing so there is no head
     space, then sealed with an aluminum crimped cap and a teflon-coated septum.
     A layer of aluminum foil is  placed between segtum and the cap to minimize
     volatile losses.   The samples are stored at 4 C until analysis.  Altern-
     atively, a 40 ml  screw-cap vial may be used with a foil-covered teflon
     disc to seal in the sample.

     Water to be analyzed for phenols, phthalates or PAH Compounds Is collected
     as part of the 24-hour composite sample or composite of grab samples used
     for general non-volatile organics analyses.

-------
3.   General Microextraction Procedure

     3.1  Place "A" ml of sample In an appropriately sized extractor (see
          Figure 1; Note 1) and sufficient salt is added to saturate the sample
          if desired.

     3.2  Add "B" microliters of solvent (pentane, hexane, diisopropylether,
          etc.) to the extractor, which is then sealed.

     3.3  Vigorously shake the extractor for 2 minutes to insure that partitioning
          of compounds of interest between water sample and organic -solvent has
          reached equilibrium.

     3.4  Place or clamp the extractor in position so that solvent rises to the
          top.

     3.5  Using a micro syringe, withdraw a sample of the solvent layer directly
          from the extractor for injection into the chromatograph or save the
          solvent layer (3.6).

     3.6  If desired, recover the solvent layer from the extractor into a small
          screw cap vial for subsequent analysis (Note 2).

4.   General Analytical Protocol

     4.1  Estimate the concentration of compounds of interest by microextracting
          a "blank" sample of wastewater, injecting 5-10 yl of extract and
          calculating the apparent concentration with an external standard and
          the following formula:

               Formula I      R01 x R03 x R04        ..
                              R02 x R05 x R06  "  yg/1

          Where:    R01 « Standard ng injected

                    R02 - Area of injected standard

                    R03 - Area of sample Injected

                    R04 • Volume of extracting solvent (yl)

                    R05 - Volume of sample extract injected (yl)

                    R06 - Sample volume extracted (ml)

Note 1:   Extractors can be of most any design or shape that allows for intimate
          contact between water and solvent during shaking.  It is desirable to
          have a small diameter neck to simplify sample removal for injection or
          storage.

Note 2:   If an emulsion forms, forcing the emulsion through a glass wool plug
          vill usually cause satisfactory separation and enable recovery of
          extracting solvent.

-------
                   A.  Silicone rubber septum
                   B.  2mm ID glass tubing
                   C.  Semlball joint
                   D.  Silicone rubber septum removed
                       when forcing solvent into
                       2"" tube
                   E.  Glass wool plug
FIGURE 1.  EXTRACTION VESSEL

(after Rhoades and Millar (2))

-------
     For example:  micro extract 40 ml ("A") of wastewater with 400 VI ("B")
     of hexane as described.   Then determine the "apparent" concentration by
     injection of 5-10 yl of the extract and calculation of the "apparent"
     concentration based on an external standard and assuming 400 VI of extract
     (Formula I).  Use the apparent concentration to plan the concentration of
     spikes and internal standard used in quantitation Formula II, Section 4.2.

     4.2  Measure two equal volumes of the same sample as the blank into each
          of two extractors.   Add an appropriate amount of IS (Internal Standard)
          to each extractor and add an appropriate amount of "spike" (based on
          the blank analysis in 4.1) to one of the extractors.  The sample to
          which only the IS has been added i& the "unspiked" sample.  The
          sample to which both the IS and the spike are added is the "spiked"
          sample.  Both samples are extracted and analyzed to collect peak area
          data as in the analysis in 4.1.  Formula II is used to calculate the
          concentration in the wastewater:
               Formula II          Rll  x   ~  -  R16
                                   R16  x  R15
                                    (R13-R16)   *

          Where:    Rll « area of unspiked sample

                    R12 » area of IS in unspiked sample

                    R13 «• area of spiked sample

                    R14 * area of IS in spiked sample

                    R.15 » ug/1 added for spike

5.   Quality Control

     Duplicate and spiked sample analyses (as described in Section 4) are
     recommended as quality control.  Each set of sample analyses is preceded
     by a method blank using organic-free water.

6.   Variations to Procedure Code #7

     6.1  Variation #1 •- Benzene, Toluene, Ethyl Benzene (Feb. 1979).
          Load 10 ml of sample into the 10-ml extraction vessel, using a syringe
          to deliver the sample through the opening at the top (remove septum D,
          Figure 1).  Next, close the top with septum D.  Carry out the entire
          loading operation either in a cold room or with the extraction vessel
          in an ice bucket.  Add 500 vl of hexane UV (Burdick and Jackson) to
          the extraction vessel through septum A.  Internal standard (o-xylene)
          and spiking solutions can also be added through septum A.  After the
          extractor has been prepared, follow the General Microextraction
          Procedure in Sections 3.3-3.6.  Analyse the extract by FID/GC using
          the following conditions:
                                     31

-------
               GC Column:     15% Carbowax 1500 on 60/80 mesh Gas Chrotn Q
                              in a 6' x A mm ID glass column.

               Carrier:       He at 20 cc/min.

               Injector:      150°C

               Detector:      100°C FID

               Temperature:   50°C (2 min), then 50°C to
                              70°C at l°C/min.

6.2  Variation 92 - Nitrobenzene (March. 1979)

     Load 40 ml of sample into a 50-ml extraction vessel, using a syringe
     to deliver the sample through the opening at the top (remove septum D,
     Figure 1).  Next, close the top with septum D.  Carry out the entire
     loading operation either in a cold room or with the extraction vessel
     in an ice bucket.  Add 500 yl of hexane UV (Burdick and Jackson) to
     the extraction vessel through septum A.  Internal Standard (nitro-
     toluene) and spiking solutions can also be added through septum A.
     After the extractor has been prepared, follow the General Micro-
     extraction Procedure in Sections 3.3-3.6.  Analyze the extract by
     GC/AFID using the following conditions:

               GC Column:     10% Carbowax 20 M + 22 KOH on Chromosorb W-
                              AW, 80/100 mesh in a 6' x A mm ID glass
                              column.

               Temperature:   160°C Isothermal

               Detector:      AFID

               Carrier:       He at 60 ml/min.

6.3  Variation #3 - Benzene - Ethyl Benzene. Toluene (May. 1979)

     Load 40 ml of sample into a 50-ml extraction vessel, using a syringe
     to deliver the sample through the opening at the top (remove septum D,
     Figure 1).  Next, close the top with septum D.  Carry out the entire
     loading operation either in a cold room or with the extraction vessel
     in an ice bucket.  Add 500 yl of hexane UV (Burdick and Jackson) to
     the extraction vessel through septum A.  Internal Standard (o-xylene)
     and spiking solutions can also be added through septum A.  After the
     extractor has been prepared, follow the General Microextraction
     Procedure in Sections 3.3-3.6.  Analyze the extract by FID/GC using
     the following conditions:
                               33

-------
               GC Column:     5% SP 1200/5% Bentone 34/10% Carbowax 1500
                              on 100/120 mesh Supelcoport, 6' x 4 mm ID
                              glass column.

               Temperature:   50°C (4 min), then 50°C to 90°C at 4°C/min.,
                              hold 90°C for 20 min.

               Detector:      200°C FID

               Injector:      180°C

               Carrier:       He at 25 ml/min.

6.4  Variation #4 - Phthalates (May.  1979)

     6.4.1     Place 100 ml of wastewater sample into a 100 ml pyrex
               volumetric glass stoppered flask.  If an internal standard
               is to be used,  it is added at this time.  The compound
               spike is also added at this time.

     6.4.2     Add 1 ml of pesticide grade hexane.

     6.4.3     Stopper and shake the flask vigorously for 2 minutes.

     6.4.4     Allow the phases to separate sufficiently to permit a
               portion of the hexane solvent to be withdrawn into a
               microliter syringe for injection into the gas chromatograph.

     6.4.5     Analyze by GC/EC with the following conditions:

               GC Column:     1.5% SP 2250/1.95% SP 2401 on 100/120 mesh
                              Supelcoport, in a 1.8 m x 6.35 mm OD glass
                              column.

               Temperature:   180°C (10 min), then 32°C/min to 220°C

               Carrier:       5% Methane/Argon at 60 ml/min

               Detector:      Electron capture at 300°C

6.5  Variation #5 - Phenols (May. 1979)

     6.5.1     Base - neutral extraction for clean-up

     6.5.1.1   Place a 200 ml aliquot of sample into a 1L separatory
               funnel equipped with a teflon stopcock.

     6.5.1.2   Check the sample pH with indicator paper.  If strongly
               acidic, it may be necessary to add NaOH pellets to raise
               the pH to 11 or above.  If the sample is neutral, add 10 N
               NaOH dropwise to reach a pH of 11 or above in the sample.
                              31

-------
          6.5.1.3   Add 25 ml of methylene chloride and shake for 1 Bin.
                    Shake cautiously and vent frequently during the first 30
                    sec. of shaking.

          6.5.1.4   Allow the organic solvent phase to separate.  Drain
                    the solvent and any emulsion into a flask and discard.

          6.5.1.5   Repeat steps 6.5.1.3 and 6.5.1.4.

          6.5.1.6   After removal of the second methylene chloride extract,
                    add 60 ml of hexane and shake for 1 minute.

          6.5.1.7   After 3-4 minutes of phase separation, rapidly drain about
                    50 ml of the aqueous phase into a flask and discard.  This
                    step is necessary to remove any residual methylene chloride
                    solvent in the stopcock and delivery tube of the separately
                    funnel.

          6.5.1.8   Drain 80 ml of sample into a 100 ml graduated cylinder.

          6.5.2     Microextraction of Sample

          6.5.2.1   Add 28 g NaCl to a 100 ml volumetric flask modified for
                    microextraction (Note 3).

          6.5.2.2   Transfer 80 ml of sample to the flask and add 1:1 RJPO,:
                    H_0 until the pH of the sample is less than 2.  Add internal
                    standard at this point if necessary.

          6.5.2.3   Add 800 yl DIE (diisopropyl ether), stopper, and shake for
                    1 minute.

          6.5.2.4   Allow phase separation prior to extract removal.  To remove
                    DIE, inject organic-free water (Millipore Super Q or
                    equivalent) slowly into the flask with an inverted U-tube
                    on top.  Collect the extract In a clean %-drain vial.  A
                    small plug of glass wool In the U-tube will help break
                    emulsions if present.
Note 3:   The 100 ml volumetric flask is modified by attaching a short length
          of glass tubing to the lower portion of the flask to allow for
          insertion of a rubber septum (as with "A" in Figure 1).  An inverted
          U-tube is made by attaching a U-shaped capillary bore glass tube to a
          ground-glass joint which fits the taper of the 100 ml volumetric
          flask.

-------
          6.5.2.5   Analyze by the following GC/FID conditions:

                    GC Column:     1% SP-1240 DA on 100/120 mesh Supelcoport
                                   in a 6' x 4 ram ID glass column with H.PO,
                                   treated glass wool plugs.

                    Carrier:       He at 100 ml/min

                    Detector:      FID

                    Temperature:   120°C (2 min), then 20°C/min to 180°C, hold
                                   20-35 min at 180°C (column can be taken to
                                   190 C, to elute sample contaminants, for
                                   short periods of time (<15 min).

                    Injector:      250°C

                    Detector:      250°C

     6.6  Variation 06 - Polynuclear Aromatic Hydrocarbons (May,  1979).

          Follow the General Microextraction Procedure for sample preparation,
          Section 3.  Analyze the hexane extract with the following GC conditions

                    GC Column:     15 meter x 0.25 mm ID SP 2100  WCOT (capillary)

                    Detector:      FID at 300°C

                    Carrier:       He at 0.75 ml/min after split

                    Split ratio:   50:1

                    Detector:      300°C

          Calculate the results as shown in Section 4.

7.   References

     1.   Henderson, J.E., Peyton, G.R., and Glaze, W.H., "A Convenient Liquid-
          Liquid Extraction Method for the Determination of Halomethanes in
          Water at the Parts-Per-Billion Level", in Identification and Analysis
          of Organic Pollutants in Water, ed. L.H. Keith, Ann Arbor Science
          Publishers, Inc., Ann Arbor, Michigan, pp 105-111 (1976).

     2.   Rhoades, J.W., and Millar,  J.D., "Gas Chromatographic Method for
          Comparative Analysis of Fruit Flowers", Agricultural and Food Chen..
          13, 5-9 (1965).

-------
Procedure Code:   8

Variations:  1-5

Title:     Purge and Trap Procedures for Analysis of Volatile
           Organic Compounds in Effluents

1.   Summary

     All the procedures included in this code are derived from the method of
     Bellar, Lichtenberg, and Eichelberger (1) and use gas stripping (purge)
     with an inert gas to remove volatile, poorly-soluble organic compounds
     from a water sample and deposit them on a suitable trapping column.  The
     trapping column is then heated and the compounds are back flushed onto
     a gas chromatographic (GC) column held at a low temperature, to deposit
     the compounds at the GC column inlet.  The chromatograph is then rapidly
     heated to its initial program temperature and the GC analysis is begun.

     Procedural variations include:  the trapping material; purge time,
     temperature, and flow; trap desorption time, temperature, and flow; the
     GC analysis parameters (column, temperature program, carrier flow); and
     the type of GC detector chosen.  Rather than assign a variation code to
     all the possible sources of variation, each procedure used by an EPA
     contractor or taken from the literature will be assigned a number, as
     follows:

     Variation           Effective
     Code No.              Date                   Description

        1                June 1977     Specific Method for Vinyl Chloride (2)
        2                June 1977     Midwest Research Method
        3a               Jan. 1979     Southwest Research Method (a)
        3b               May 1979      Southwest Research Method (b)
        4                Feb. 1979     Method for Acrylonitrile
        5                May 1979      Method for Acrolein and Acrylonitrile

2.   Sample Collection and Compositing

     Samples are collected in 40 ml serum vials, filled to overflowing so
     that there is no head space, then sealed with an aluminum crimp cap and
     a teflon-coated septum.  A layer of aluminum foil is placed between the
     septum and the cap to minimize volatile losses.  Alternatively, the
     samples may be collected in 40 ml vials with screw caps, using teflon
     discs covered with aluminum foil to seal the sample head-space free.
     The samples are stored at 4 C until analysis.  For compositing, the cold
     samples are combined in a chilled flask (4 C);  then an aliquot is
     transferred to the purging device.

3.   Vinyl Chloride Procedure
     Variation #1

     This method is described in detail in the paper by Bellar, Lichtenberg,
     and Eichelberger (1).  It uses a silica gel or carbosieve-B trap, designed

-------
     to retain vinyl chloride.   The method is hence compound-specific and should
     be used in verification sampling when vinyl chloride is  the only "purgeable"
     compound present in the sample.  ' If other purgeable compounds  are present,
     a liquid-liquid extraction method should be tried,  or Variations //2  and #3
     of the purge and trap procedure may be used if the  sample matrix makes
     liquid-liquid extraction ineffective.

     Envirodyne Engineers, Inc., Parameters for Vinyl Chloride

         Purge:  12 min.  at 40 ml/min. He, ambient temperature, 5 ml sample.
                      »

         Trap:  4.5 in X 2 mm ID, Carbosieve-B, 60/80 mesh.

         Desorb:  4 min.  at 180°C, 30 ml/min He.

         GC Column:  8 ft. X 2 mm ID stainless steel with 0.2% Carbowax
                     1500 on Carbopack C, 60/80 mesh.

         Temperature:  Ambient during desorption, then heat to 60 C, hold
                       4 min., and program to 170  at 8  /min.

         Detector:  Hall electrolytic conductivity in halide  mode or mass
                    spectrometer.

         Carrier:   He at 30 ml/min.

4.   Variation 92 (Midwest Research)

         Purge:  8 min.  at 40 ml/min. N_, ambient temperature, 1-5  ml
                 sample.

         Trap:   10 in.  x 0.125 in. OD  x 105 in. ID stainless steel, Tenax
                 GC, 60/80 mesh.

         Desorb: 4 min.  at 180°C

         GC Column:  8 ft. x 0.125 in. OD stainless steel with 0.2% Carbowax
                     1500 on Carbopack C, 60/80 mesh.

         GC Temperature:  4 min. at 40°C, 8°/min. to 160°C, hold 5  min.

         Detector:  FID and/or Hall electrolytic conductivity (halide mode)

         Quantitation:  Use area ratio to bromochloromethane as the internal
                        standard.

5.   Variation #3(a)  (Southwest Research)

         Purge:  12 min. at 40 ml/min. He, ambient temperature, 5 ml sample.

         Trap:   6 in. x 4 mm ID, Tenax GC, 60/80 mesh.

-------
    Desorb:  A min. at 180°C, 30 ml/min. He

    GC Column:  8 ft. x 2 mm ID nickel, with 0.2% Carbowax 1500 on
                Carbopack C, 60/80 mesh.

    Temperature:  Ambient during desorption: then heat to 60°C, hold 4 min. ,
                  and program to 170 C at 8 /min.

    Detector:  Hall electrolytic conductivity in halide mode.

    Carrier:  Helium at 30 ml/min

    Quantitation:  Concentrations of analytes are determined as follows:


               *S " ^J " AAS

               AAS
               -££.. a
               ngs
               Tp
               vp


Where:
               AS  •  Area of peak in spiked run

               A..  •  Area of peak in unspiked run

               A._ •  Area of added epike
                AS
               o  •  Constant, area unita per ng
               M..  •  mass (ng) in unaplked peak

               Vp  •  volume purged (ml)

               C.  »  Concentration, ppb


Variation //3(b) (Southwest Research)

Same as Variation #3(a)  except:

    Purge Flow:  30 ml/min.  He

    Trap:  6 in.  X 4 mm ID, with 5 in.  of Tenax GC,  60/80 mesh at the
           inlet, and 1 in.  of silica gel 35/60 mesh, placed after the
           Tenax in the trap  flow direction.

    Desorb Flow:   22 ml/min.  He

    GC Carrier Flow:  22 ml/min. He

-------
6.   Variation #4 (Method for Acrylonitrile)  (Source:  EEI)*

         Purge:  12 min. at 40 ml/min., at 60°C, with Helium; 25 ml sample
                 volume purged.

         Trap:  6 in. x 4 mm ID stainless steel, Tenax GC, 60/80 mesh.

         Desorb:  4 min. at 180°C, 30 ml/min.

         GC Column:  6' x 1/4" OD glass column with Chromosorb 102, 60/80 mesh.

         GC Temperature:  100 C isothermal.

         Injector:  200°C.

         Detector:  FID at 200°C.

         Carrier:   Helium at 30 ml/min.

     * See Procedure Code #1 for direct aqueous injection method for acrylonitrile.

7.   Variation #5 (Method for Acrolein and Acrolonitrile)  (Source: EEI)

         Purge:  12 min. at 40 ml/min He, at 60°C, 25-40 ml sample.  Composite
                 samples at 4 C, if necessary, before analysis.

         Trap:  2 mm ID X 4.5 in. stainless steel, with 60/80 mesh Tenax GC.

         Desorb:  7 min. x 200°C, 30 ml/min. He.

         GC Column:  Chromosorb 102 (60/80 mesh) in a 6' x 1/4" glass column.

         Temperature:  110°C isothermal.

         Detector:  FID at 250°C.

         Carrier:  He at 35 ml/min.

8.   References

     1.  Bellar, T. A., Lichtenberg, J. J., and Eichelberger, J. W.,
         "Determination of Vinyl Chloride at mg/1 Level in Water by Gas Chro-
         ma to graphy," EnvJ^pjuS£i_:__andT^chnoL:_jL_10, 926-930 (1976).

-------
Procedure Code: 9

Title:  Method for Polychlorinated Biphenyls (PCB's) in Industrial Effluents

1.   Scope and Application

     1.1  This method covers the determination of certain polychlorinated
          biphenyl (PCB) mixtures, including: Aroclors 1221, 1232, 1242, 1248,
          1254, 1260 and 1016.

     1.2  The method is an extension of Procedure //3, the EPA method for
          organochlorine pesticides and phthalate esters in industrial effluents
          (1).  It is designed so that determination of PCB's, organochlorine
          pesticides, and phthalate esters may be made on the same sample.

     1.3  The limit of detection is approximately I/ g/1 for each Aroclor
          mixture.

2.   Summary

     2.1  The PCB's and the organochlorine pesticides are coextracted by
          liquid-liquid extraction and, insofar as possible, the two classes
          of separated from one another prior to gas chromatographic determination.
          combination of the standard Florisil column cleanup procedure and a
          gel microcolumn separation procedure (2,3) is employed.  Identification
          is made from gas chromatographic patterns obtained through the use
          of two or more unlike columns.  Detection and measurement is accomplished
          by using an electron-capture detector.  Techniques for confirming
          qualitative identification are suggested.

3.   Interferences

     3.1  Solvents, reagents, glassware, and other sample processing hardware
          may yield discrete artifacts and/or elevated baselines causing mis-
          interpretation of gas chromatograms.   All these materials must be
          demonstrated to be free of interferences under the conditions of the
          analysis.  Specific selection of reagents and purification of solvents
          by distillation in all-glass systems may be required.  Refer to (4),
          Part I. Sections 1.4 and 1.5.

     3.2  The interferences in industrial effluents are high and varied, and
          pose great-difficulty in obtaining accurate and precise measurement
          of PCB's.  Separation and cleanup procedures are generally required
          to eliminate these interferences; however, such techniques may
          result in the loss of certain organochlorine compounds.  For this
          reason, great care should be exercised in the selection and use of
          methods for eliminating or minimizing interferences.   It is not
          possible to describe procedures for overcoming all the interferences
          that may be encountered in industrial wastes.

-------
     3.3  Phthalate esters, certain organophosphorus pesticides, and elemental
          sulfur will interfere when using electron capture for detection.
          Organochlorine pesticides and other halogenated compounds also
          constitute interferences in the determination of PCB's.  Most of
          these are separated by the method described herein.  However, certain
          compounds, if present in the sample, will occur with the PCB's.
          Included are: sulfur, Heptachlor, aldrin, DDE, technical chlordane,
          mirex, and to some extent o,p'-DDT and p,p'-DDT.

A.   Apparatus and Materials

     A.I  Gas Chromatograph - Equipped with glass-lined injection port.

     4.2  Detector - Electron Capture

     4.3  Recorder - Potentiometric strip chart (10 in.) compatible with
          detector system.

     4.4  Gas Chromatographic Column Materials:

          4.4.1     Tubing - Pyrex (180 cm long x 4 mm ID/or 6' x 2 mm ID)

          4.4.2     Glass Wool - Silanized

          4.4.3     Solid Support* - Gas-Chrom Q (100/120 mesh)

          4.4.4     Liquid Phases* - Expressed as weight percent coated on solid
                    Support:

                    4.4.4.1  SE-30 or OV-1, 3%

                    4.4.4.2  OV-17, 1.57. + QF-1, 1.95%

     4.5  Kuderna-Danish  (K-D) Glassware (Kontes)

          4.5.1     Snyder Columns - Three ball (macro)

          4.5.2     Evaporate Flasks - 500 ml

          4.5.3     Receiver Ampuls - 10 ml graduated

          4.5.4     Ampul stoppers


          *See also Section 13.

-------
     4.6  Chromatographic Column - Chromaflex  (400 mm long x  19 mm  ID) with
          coarse fritted plate on bottom and Teflon stopcock;  250 ml reservoir
          bulb at top of column with flared out funnel shape  at top of bulb -
          a special order (Kontes K-420540-9011).

     4.7  Chromatographic Column - Pyrex (approximately 400 mm long x 20 nun ID)
          with a coarse fritted plate on bottom.

     4.8  Micro Column Pyrex - constructed according to Figure 1.

     4.9  Capillary pipets disposable (5-3/4 in.) with rubber  bulb.  (Scientific
          Products P5205-1).

     4.10 Low-pressure regulator - 0 to 5 psig - with low-flow needle valve
          (See Figure 1, Matheson Model 70).

     4.11 Beaker - 100 ml.

     4.12 Micro syringes - 10, 25, 50 and 100 jjl.

     4.13 Separatory Funnels - 125 ml,  1000 ml, and 2000 ml with Teflon stopcocks,

     4.14 Graduated Cylinders - 100 ml,  250 ml.

     4.15 Blender - High speed, glass or stainless cup.

     4.16 Florisil - PR Grade (60/100 mesh); purchase activated at 125°F
          and store in the dark in glass containers with glass stoppers or
          foil-lined screw caps.  Before use, activate each batch overnight at
          130° in foil-covered glass container.  Determine lauric-acid value
          (See Section 14).

     4.17 Silica gel - Davison code 950-08-08-226 (60/200 mesh).

     4.18 Glass Wool - Hexane extracted.

     4.19 Centrifuge Tubes - Pyrex calibrated (15 ml).

5.   Reagents, Solvents, and Standards

     5.1  Ferrous Sulfate - (ACS) 30% solution in distilled water.

     5.2  Potassium Iodide - (ACS) 10% solution in distilled water.

     5.3  Sodium Chloride - (ACS) Saturated solution (pre-rinse NaCl with
          hexane) in distilled water.

     5.4  Sodium Hydroxide - (ACS) 10 N in distilled water.

     5.5  Sodium Sulfate - (ACS) Granular,  anhydrous,  conditioned for 4 hours
          at 400° C.

-------
     5.6  Sulfuric Acid - (ACS) Mix equal volumes of cone. H?SO  with distilled
          water.                                            2  4

     5.7  Diethyl Ether - Nanograde, redistilled in glass, if necessary.

          5.7.1     Must contain 22 alcohol and be free of peroxides by
                    following test:  to 10 ml of ether in glass-stoppered
                    cylinder previously rinsed with ether, add one ml of freshly
                    prepared 10% KI solution.  Shake and let stand one minute.
                    No yellow color should be observed in either layer.

          5.7.2     Decompose ether peroxides by adding 40g of 30% ferrous
                    sulfate solution to each liter of solvent.  CAUTION: Reaction
                    may be vigorous if the solvent contains a high concentration
                    of peroxides.

          5.7.3     Distill deperoxidized ether in glass, and add 2% ethanol.

     5.8  n-Hexane - Pesticide quality (NOT MIXED HEXANES).

     5.9  Acetonitrile, Hexane. Methanol, Methylene Chloride, Petroleum Ether
          (Boiling range 30-60 C) - pesticide quality,  redistilled in glass if
          necessary.

     5.10 Standards - Aroclors 1221, 1232, 1242, 1248,  1254,  1260, and 1016.

     5.11 Anti-static Solution - STATNUL, Daystrom, Inc., Weston Instrument
          Division, Newark,  New Jersey  95212.

6.   Calibration

     6.1  Gas chromatographic operating conditions are considered acceptable
          when the response to dicapthon is at least 50% of full scale when
          0.06 ng or less is injected for electron-capture detection. For all
          quantitative measurements, the detector must be operated within its
          linear response range, and the detector noise level should be less
          than 2% of full scale.

     6.2  Standards are injected frequently as a check on the stability of
          operating conditions, detector, and column.  Example chromatograr.s,
          shown in Figures 3 through 8, provide reference operating conditions.

'•   Quality Control

     7.1  Duplicate and spiked sample analyses are recommended as a quality
          control check.  When the routine occurrence of a pollution parameter
          is observed, quality control charts are also recommended (5).
                                 T-

-------
     7.2  Each tine a set of samples is extracted, a method blank  is  determined
          on a volume of distilled water equal to that used to dilute the sample.

8.   Sample Preparat ion

     8.1  Blend the sample if suspended matter is present, and adjust pH to near
          neutral (pH 6.5-7.5) with 50% sulfuric acid or 10 N sodium  hydroxide.

     8.2  For sensitivity requirement of 1 yg/1, when using microcoulometric
          or electrolytic conductivity methods for detection take  1000 ml of
          sample for analysis.  If interferences pose no problem,  the sensitivity
          of the electron-capture detector should permit a little  as  100 ml of
          sample to be used.  Background information on the extent and nature
          of interferences will assist the analyst in choosing the required sample
          size and preferred detector.

     8.3  Quantitatively transfer the proper aliquot into a two-liter  separatory
          funnel, and dilute to one liter.

9.   Extraction

     9.1  Add 60 ml of 15% methylene chloride in hexane (v:v) to the  sample
          in the separatory funnel, and shake vigorously for two minutes.

     9.2  Allow the mixed solvent to separate from the sample;, then  draw the
          water into a one-liter Erlenmeyer flask.  Pour the organic  layer* into
          a 100 ml beaker, and then pass it through a column containing 3-4
          inches of anhydrous sodium sulfate and collect it in a 500 ml K-D
          flask equipped with a 10 ml ampule.  Return the water phase  to the
          separatory funnel.  Rinse the Erlenmeyer flask with a second 60 ml
          volume of solvent; add the solvent to the separatory funnel and
          complete the extraction procedure a second time.   Perform a  third
          extraction in the same manner.

     9.3  Concentrate the extract to 6-10 ml in the K-D evaporator on a hot
          water bath.

     9.4  Qualitatively analyze the sample by gas chromatography with an
          electron-capture detector.  From the response obtained, decide:

          a.   If there are any organochlorine pesticides present.

          b.   If there are any PCB's present,

          c.   If there is a combination of a and b,

          d.   If elemental sulfur is present,

          e.   If the response is too complex to determine a,b, or c.

-------
               f.   If no response, concentrate to 1.0 ml or less, as required,
                    according to EPA Method (A), pg. 28, and repeat the analysis
                    looking for a, b, c, d, and e.  Samples containing Aroclors
                    with a low percentage of chlorine, e.g., 1221 and 1232, may
                    require this concentration in order to achieve the detection
                    limit of 1 yg/1.  Trace quantities of PCB's are often
                    masked by background which usually occur in the samples.

     9.5  If condition a_ exists, quantitatively determine the organochlorine
          pesticides according to Procedure #3, "Method for Organochlorine
          Pesticides and Phthalate Esters in Industrial Effluents."

     9.6  If condition b_ exists, PCB's only are present, and no further separa-
          tion or cleanup is necessary.  Quantitatively determine the PCB's in
          accordance with Section 11.

     9.7  If condition £ exists, compare peaks obtained from the sample to
          those of standard Aroclors, and make a judgment as to which Aroclors
          may be present.  To separate the PCB's from the organochlorine
          pesticides, continue as outlined in 10.4.

     9.8  If condition ^ exists, separate the sulfur from the sample using
          the method outlined in (10.3) followed by the method in (10.5).

     9.9  If condition je exists, then the following macro cleanup and separation
          procedures (10.2 and 10.3) should be employed and, if necessary,
          followed by the micro separation procedures (10.4 and 10.5).

10.  Cleanup and Separation Procedures

     10.1 Interferences in the form of distinct peaks and/or high background
          in the initial gas chromatographic analysis, as well as the physical
          characteristics of the extract (color, cloudiness, viscosity) and
          background knowledge of the sample,  will indicate whether cleanup is
          required.  When these interfere with measurement of the pesticides or
          affect column life or detector sensitivity, proceed as directed in
          the following paragraphs.

     10.2 Acetonitrile Partition - This procedure is used to remove fats and
          oils from the sample extracts.  It should be noted that not all
          pesticides are quantitatively recovered by this procedure.   The
          analyst must be aware of this and demonstrate the efficiency of the
          partitioning for the compounds of interest.

     10.2.1    Quantitatively transfer the previously concentrated extract
               to a 125 ml separatory funnel with enough hexane to bring the
               final volume to 15 ml.  Extract the sample four times by shaking
               vigorously for one minute with 30-ml portions of hexane-saturated
               acetonitrile.

-------
     10.2.2    Combine and transfer the acetonitrile phases  to a one-
               liter separatory funnel, and add 650 ml of distilled water
               and 40 ml of saturated sodium chloride solution.  Mix
               thoroughly for 30-35 seconds.  Extract with two 100-ml
               portions of hexane by vigorously shaking about 15 minutes.

     10.2.3    Combine the hexane extracts in a one-liter separatory funnel,
               and wash with two 100 ml portions of distilled water.
               Discard the water layer and pour the hexane layer through a
               3-A inch anhydrous sodium sulfate column into a 500 ml 
-------
                                      6Z Eluate

               Aldrin         DDT                    Pentachloro-
               BHC            Heptachlor              nitrobenzene
               Chlordane      Heptachlor Epoxide     Strobane
               DDD            Lindane                Toxaphene
               DDE            Methoxychlor           Trifluralin
                              Mirex                  PCB's

                   152 Eluate           50% Eluate

                   Endosulfan I         Endosulfan II
                   Endrin               Captan
                   Dieldrin
                   Dichloran
                   Phthalate esters

               Certain thiophosphate pesticides will occur in each of the
               foregoing fractions and in the 100% fraction.  For additional
               information regarding eluate composition, refer to the FDA
               Pesticide Analytical Manual (6).

     10.3.4    Concentrate the eluates to 6-10 ml in the K-D evaporator
               in a hot-water bath.

     10.3.5    Analyze by gas chromatography.

10.4 Silica Gel Microcolumn Separation Procedure (7)

     10.4.1    Activation for Silica Gel

          10.4.1.1  Place about 20 em of silica gel in a 100 ml beaker.
                    Activate at 180 C for approximately 16 hours.  Transfer
                    the silica gel to a 100 ml glass stoppered bottle.
                    When cool, cover with about 35 ml of 0.50% diethy1
                    ether in benzene (volume:volume).  Keep bottle well
                    sealed.  If silica gel collects on the ground glass
                    surfaces, wash off with the solvent before resealing.
                    Always maintain an excess of the mixed solvent in
                    bottle (approximately 1/2 in. above silica gel).
                    Silica gel can be effectively stored in this manner
                    for several days.

     10.4.2    Preparation of the Chromatographic Column

          10.4.2.1  Pack the lower 2-mm ID section of the microcolumn
                    with glass wool.  Permanently mark the column 120 mm
                    above the glass wool.  Using a clean rubber bulb from
                    a disposable pipet, seal the lower end of the micro-
                    column.  Fill the microcolumn with 0.50% ether in
                    benzene  (v:v) to the bottom of the 10/30 Joint (Figure
                    1).  Using a disposable capillary pipet, transfer
                    several aliquots of the silica gel slurry into the
                    microcolumn.  After approximately 1 cm of silica gel

-------
          collects in the bottom of the nlcrocolumn, remove the
          rubber bulb seal, and tap the column to insure  that
          the silica gel settles uniformly. Carefully pack the
          column until the silica gel reaches the 120 + 2 mm
          mark.  Be sure that there are no air bubbles in the
          column.  Add about 10 mm of sodium sulfate to the top
          of the silica gel.  Under low-humidity conditions, the
          silica gel may coat the sides of the column and not
          settle properly.  This can be .minimized by wiping the
          outside of the column with an anti-static solution.

10,.4.2.2  Deactivation of the Silica Gel

          a.   Pill the microcolumn to the base of the 10/30
               joint with the 0.50% ether-benzene mixture,
               assemble reservoir (using spring clamps), and
               fill with approximately 15 ml of the 0.50% ether-
               benzene mixture.  Attach the air pressure device
               (using spring clamps) and adjust the elution rate
               to approximately 1 ml/tnin. with the air-pressure
               control.  'Release the air pressure and detach the
               reservoir just as the last of the solvent enters
               the sodium sulfate.  Fill the column with n-
               hexane (not mixed hexanes) to the base of the
               10/30 fitting.  Evaporate all residual benzene
               from the reservoir, assemble the reservoir section,
               and fill with 5 ml of n-hexane.  Apply air pressure
               and adjust the flow to 1 ml/min.  (The n-hexane
               flows slightly faster than the benzene).  Release
               the air pressure, and remove the reservoir just
               as the n-hexane enters the sodium sulfate.  The
               column is now ready for use.

          b.   Pipet a 1.0 ml aliquot of the concentrated sample
               extract (previously reduced to a total volume of
               2.0 ml) onto the column.  As the last of the
               sample passes into the sodium sulfate layer,
               rinse down the internal wall of the column twice
               with 0.25 ml of n-hexane.  Then assemble the
               upper section of the column.  As the last of the
               n-hexane rinse reaches the surface of the sodium
               sulfate, add enough n-hexane (volume predetermined,
               see 10.4.3) to just elute all the PCB's present
               in the sample.  Apply air pressure and adjust
               until the flow is 1 ml/min.  Collect the desired
               volume of eluate (predetermined, see 10.4.3)
               in an accurately calibrated ampule.  As the last
               of the n-hexane reaches the surface of the sodium
               sulfate, release the air pressure and change the
               collection ampule.
                _L-

-------
               c.   Fill the column with 0.50% diethyl ether in
                    benzene, again apply air pressure, and adjust
                    flow to 1 ml/min.  Collect the eluate until all
                    the organochlorlne pesticides of Interest have
                    been eluted (volume predetermined, tee 10.4.3).

               d.   Analyze the eluates by gas chromatography.

10.4.3    Determination of Elution Volumes

     10.4.3.1  The elutlon volumes for the PCB's and the pesticides •
               depend upon a number of factors which are difficult to
               control.  These include variation in:

               a.   Mesh size- of the silica gel

               b.   Adsorption properties of the silica gel

               c.   Polar contaminants present in the eluting solvent

               d.   Polar materials present in the sample and sample
                    solvent

               e.   The dimensions of the microcolumns

               Therefore, the optimum elution volume must be experi-
               mentally determined each time a factor is changed.  To
               determine the elution volumes, add standard mixtures
               of Aroclors and pesticides to the column and serially
               collect 1-ml elution volumes.  Analyze the individual
               eluates by gas chromatography and determine the cut-
               off volume for n-hexane and for ether-benzene.  Figure
               2 shows the retention order of the various PCS compon-
               ents and of the pesticides. Using this information,
               prepare the mixtures required for calibration of the
               microcolumn.

     10.4.3.2  In determining the volume of hexane required to elute
               the PCB's, the sample volume  (1 ml) and the volume of
               n-hexane used to rinse the column wall must be consid-
               ered.  Thus, if it is determined that a 10.0 ml elution
               volume is required to elute the PCB's, the volume of
               hexane to be added, in addition to the sample volume
               but including the rinse volume, should be 9.5 ml.

     10.4.3.3  Figure 2 shows that as the average chlorine content
               of a PCB mixture decreases the solvent volume for
               complete elution increases.  Qualitative determination
               (9.4) indicates which Aroclors are present and provides
               the basis for selection of the ideal elution volume.

-------
                     This helps to minimize the quantity of organochlorine
                    pesticides which will elute along with the low-percent-
                    chlorine PCB's and insures the most efficient separations
                    possible for accurate analysis.

          10.4.3.4  For critical analysis where the PCB's and pesticides
                    are not separated completely, the column should
                    be accurately calibrated (according to 10.4.3.1)
                    to determine the percent of material of interest that
                    elutes in each fraction.  Then flush the column with
                    an additional IS ml of 0.50% ether in benzene followed
                    by 5 ml of n-hexane, and use this reconditioned column
                    for the sample separation.  Using this technique one
                    can accurately predict the amount (%) of materials in
                    each microcolumn fraction.

10.5 Microcolumn Separation of Sulfur, PCB's, and Pesticides

     10.5.1    See procedure for preparation and packing microcolumn
               in PCB analysis section (10.4.1 and 10.4.2).

     10.5.2    Microcolumn Calibration

          10.5.2.1  Calibrate the microcolumn for sulfur and PCB separation
                    by collecting 1.0 ml fractions and analyzing them by
                    gas chromatography to determine the following:

                    a.   The fraction with the first eluting PCB's (those
                         present in 1260)

                    b.   The fraction with the last eluting PCB's
                         (those present in 1221)

                    c.   The elution volume for sulfur

                    d.   The elution volume for the pesticides of interest
                         in the 0.50% ether-benzene fraction

                    From these data, determine the following:

                    e.   The eluting volume containing only sulfur
                         (Fraction I)

                    f.   The eluting volume containing the last of
                         the sulfur and the early-eluting PCB's (Fraction
                         ID

                    g.   The eluting volume containing the remaining
                         PCB's (Fraction 111)

                    h.   The ether-benz.e*e. eluting volume containing
                         the pesticides oi interest (Fraction iv)


-------
          10.5.3    Separation Procedure

               10.5.3.1  Carefully concentrate the 6% eluate from  the florisil
                         column to 2.0 ml in the graduated ampule  on a warm
                         water bath.

               10.5.3.2  Place 1.0 ml (50%) of the concentrate into the
                         microcolumn with a 1 ml pipet.  Be careful not  to get
                         any sulfur crystals into the pipet.

               10.5.3.3  Collect Fractions I and II in calibrated  centrifuge
                         tubes.

                         Collect Fractions III and IV in calibrated ground-
                         glass-stoppered ampules.

               10.5.3.4  Sulfur Removal (9) - Add 1 to 2 drops of mercury
                         to Fraction II stopper and place on a wrist-action
                         shaker.  A black precipitate indicates the presence of
                         sulfur.  After approximately 20 minutes the mercury
                         may become entirely reacted or deactivated by the
                         precipitate.  The sample should be quantitatively
                         transferred to a clean centrifuge tube and additional
                         mercury added.   When crystals are present in the
                         sample, three treatments may be necessary to remove
                         all the sulfur.  After all the sulfur has been removed
                         from Fraction II (check using gas chromatography)
                         combine Fractions II and III.  Adjust the volume to 10
                         ml and analyze by gas chromatography.  Be sure no
                         mercury is transferred to the combined Fractions II
                         and III,  since it can react with certain pesticides.

                         By combining Fractions II and III (if PCB's are
                         present), it is possible to identify the Aroclor(s)
                         present,  and a quantitative analysis can be performed
                         accordingly.  Fraction I can be discarded since it
                         contains only the bulk of the sulfur.  Analyze Fractions
                         III and IV for the PCB's and pesticides.

                         If DDT and its homologs (aldrin, heptachlor,  and/or
                         technical chlordane) are present along with the PCB's,
                         an additional microcolumn separation can be performed;
                         this may help to further separate the PCB's from the
                         pesticides (See 10.4)

11.  Quantitative Determination

     11.1 Measure the volume of n-hexane eluate containing the PCB's,  and
          inject 1 to 5 pi into the gas chromatograph.  If necessary,  adjust
          the volume of the eluate to give linear response to the electron-
          capture detector.

-------
11.2 Calculations

     11.2.1    When a single Aroclor IB present, compare quantitative
               Aroclor reference standards (e.g., 1242, 1260) to the
               unknown.  Measure and sum the areas of the unknown and the
               reference Aroclor, and calculate the result as follows:

                                          (A)  (B)  (V )
                       Micrograra/liter -  	-rr-r	rrr-r—   X N
                                            *V  * s}
                       A m ng °^ Standard Injected   m ng_
                           Z of Standard Peak Areas    mm
                                                       2
                       B « £ of Sample Peak Areas * (mm )

                       V  • Volume of sample injected (yl)

                       V  « Volume of Extract (yl) from which sample
                            is injected into gas chromatograph

                       V  » Volume of water sample extracted (ml)

                       N « 2 when microcolumn is used
                           1 when microcolumn is not used

                       Peak Area - Peak height (mm)  x Peak Width at 1/2
                                   height, or calculated by an electronic
                                   integrator)

     11.2.2    For complex situations,  use the calibration method
               described in the following paragraphs.  Small variations in
               conponents between different Aroclor batches make it
               necessary to obtain samples of several specific Aroclors.
               These are available from commercial chromatography supply
               companies.

          11.2.2.1  Using the OV-1 column, chromatograph a known
                    quantity of each Aroclor reference standard. Also
                    chromatograph a sample of p,p'-DDE.  Suggested concen-
                    tration of each standard is 0.1  ng/1 for the Aroclors
                    and 0.02 ng/1 for the p,p'-DDE.

          11.2.2.2  Determine the relative retention time (RRT) of
                    each PCB peak in the resulting chromatograms using
                    p,p'-DDE as 100.  See Figures 3 through 5.
                          RT  x  100
                    RRT -
                            RTDDE
                    RRT • Relative Retention Time

                    RT - Retention time of peak of interest
                    RT
                      DDE * Retention time of p,p'-DDE

-------
          Retention time is measured as that distance in
          nun between the first appearance of the solvent peak
          and the maximum for the compound.

11.2.2.3  To calibrate the instrument for each PCS, measure
          the area of each peak.

          Area - Peak height (mm) x Peak width at 1/2 height.
          Using Tables 1 through 6, obtain the proper mean
          weight factor; then determine the response factor
          g/mm
                   (ng.)  (mean weight percent)
          ng/mm  -
          ng. » ng of Aroclor Standard Injected

          Mean weight percent » obtained from Tables 1
          through 6.

11.2.2.4  Calculate the RRT value and the area for each PCB
          peak in the sample chromatogram.  Compare the sample
          chromatogram to those obtained for each reference
          Aroclor standard.  If it is apparent that the PCB
          peaks present are due to only one Aroclor, then
          calculate the concentration of each PCB using the
          following formula:
                        2
          ng PCB - ng/mm  x Area
                                 2
          Where:  Area - Area (mm ) of sample peak

                  ng/mm  « Response factor for that peak measured.

          Then add the nanograms of PCB's present in the
          injection to get the total number of nanograms of
          PCB's present.  Use the following formula to calculate
          the concentration of PCB's in the sample:

                            (£ng)   (V£)
          micrograms/Liter »  (v ) (v )  x N

          V  « volume of water extracted  (ml)

          V  » volume of extract (ul)

          V. • volume of sample injected  (yl)

          2ng - sum of all the PCB's in nanograms for  that
          Aroclor identified

          N » 2 when mlcrocoluom is used

          N • 1 when microcolunm is not used

-------
                         The value can then be reported as micrograms/liter
                         PCS's reported as the Aroclor.  For samples containing
                         more than one Aroclor, use Figure 9 chronatogram
                         divisional flow chart to assign a proper response
                         factor to each peak and also identify the "most likely"
                         Aroclors present.  Calculate the ng of each PCB isomer
                         present, and sum them according to the divisional flow
                         chart.  Using the formula above, calculate the concen-
                         tration of the various Aroclors present in the sample.
12.  Reporting Results
     Report results in micrograms per liter without correction for recovery
     data.  When duplicate and spiked samples are analyzed, all data obtained
     should be reported.

13.  Variations to Procedure #9

     Variation      Effective
        No.           Date              Description

        1           May 1978            GC Column:   7% OV17 + 9% OV210 on
                                                     80/100 Chromosorb W in a
                                                     a 6 ft. x 2 mm ID glass
                                                     column
                                        Temperature:  190 C isothermal
                                        Injection:    27§ic
                                        Detector:     Ni   at 325 C
                                        Carrier:      Ar/Me at 60 ml/min

14.  Standardization of Florisil Column by Weight Adjustment Based on Adsorption
     of Lauric Acid.

     14.1 A rapid method for determining adsorptive capacity of Florisil is
          based on adsorption of lauric acid from hexa'ne solution (6, 10). An
          excess of lauric acid is used, and the amount not adsorbed is measured
          by alkali titration.  Weight of lauric acid adsorbed is used to
          calculate, by simple proportion, equivalent quantities of florisil
          for batches having different adsorptive capacities.

     14.2 Apparatus

          14.2.2    Buret. — 25 ml graduations.

          14.2.2    Erlenmeyer flasks. — 125 ml narrow mouth and 25 ml,
                    glass stoppered.

          14.2.3    Pipet. — 10 and 20 ml transfer.

          14.2.4    Volumetric flasks. — 500 ml.

-------
14.3 Reagents and Solvents

     14.3.1    Alcohol, ethyl. — USP or absolute, neutralized to
               phenolphthalein.

     14.3.2    Hexane, — Distilled from all-glass apparatus,

     14.3.3    Laurie acid. — Purified, CP.

14.3.4    Laurie acid solution. — Transfer 10.000 g lauric acid to 500
          ml volumetric flask, dissolve in benzene, and dilute to 500 ml
          (ml * 20 mg).

14.3.5    Phenolphthalein Indicator. — Dissolve 1 g in alcohol and dilute
          to 100 ml.

14.3.6    Sodium hydroxide. — Dissolve 20 g NaOH (pellets, reagent grade)
          in water and dilute to 500 ml (IN).  Dilute 25 ml IN NaOH to 500
          ml with water (0.05N).  Standardize as follows:  Weigh 100-200
          mg lauric acid into 125 ml Erlenmeyer flask.  Add 50 ml neutral-
          ized ethyl alcohol and 3 drops phenolphthalein indicator; titrate
          to permanent end point.  Calculate mg lauric acid/ml 0.05 JJ NaOH
          (about 10 mg/ml).

14.4 Procedure

     14.4.1    Transfer 2.000 g Florisil to 25 ml glass-stoppered
               Erlenmeyer flasks.  Cover loosely with aluminum foil and
               heat overnight at 130 C.  Stopper, and cool to room
               temperature; add 20.0 ml lauric acid solution (400 mg),
               stopper, and shake occasionally for 15 min.  Let adsorbent
               settle and pipet 10.0 ml of supernatant into 125 ml
               Erlenmeyer flask.  Avoid inclusion of any Florisil.

     14.4.2    Add 50 ml neutral alcohol and 3 drops indicator solution;
               titrate with 0.051? to a permanent end point.

14.5 Calculation of Lauric Acid Value and Adjustment of Column Weight

     14.5.1    Calculate amount of lauric acid adsorbed on Florisil
               as follows:

               Lauric Acid value - mg lauric acid/g Florisil - 200 -
               (ml required for titration x mg lauric acid/ml 0.05N NaOH).

     14.5.2    To obtain an equivalent quantity of any batch of Florisil,
               divide 110 by lauric acid value for that batch, and multiply
               by 20 g.  Verify proper elution of pesticides by 13.6.

14.6 Test for Proper Elution Pattern and Recovery of Pesticides.
     Prepare a test mixture containing aldrin, heptachlor epoxide, p,p'-
     DDE, dieldrin, Parathion, and malathion.  Dieldrin and Parathion
     should elute in the 15% eluate and all but a trace of malathion in
     the 50% eluate and the others in the 6% eluate.

-------
15.  References

(1)  "Method for Organochlorine Pesticides in Industrial Effluents," U.S.
     Environmental Protection Agency, National Environmental Research Center,
     Analytical Quality Control Laboratory, Cincinnati, Ohio 45268, 1973.

(2)  Leoni, V., "The Separation of Fifty Pesticides and Related Compounds
     and Polychlorinated Biphenyls into Four Groups by Silica Gel Micro-column
     Chromatography," Journal of Chromatography, 62. 63 (1971).

(3)  McClure, V. E., "Precisely Deactivated Adsorbents Applied to the
     Separation of Chlorinated Hydrocarbons," Journal of Chromatoeraphy,
     20^168(1972).

(A)  "Methods for Organic Pesticides in Water and Wastewater," U.S. Environ-
     mental Protection Agency, National Environmental Research Center,
     Analytical Quality Control Laboratory, Cincinnati, Ohio 45268, 1971.

(5)  "Handbook for Analytical Quality Control in Water and Wastewater
     Laboratories," Chapter 6, Section 6.4, U.S. Environmental Protection
     Agency, National Environmental Research Center, Analytical Quality
     Control Laboratory, Cincinnati,  Ohio 45268, 1972.

(6)  "Pesticide Analytical Manual," U.S.  Dept.  of Health,  Education, and
     Welfare, Food and Drug Administration, Washington, D.C.

(7)  Bellar, T. A.  and Lichtenberg, J.J., "Method for the  Determination
     of Polychlorinated Biphenyls in Water and  Sediment,"  U.S.  Environmental
     Protection Agency, National Environmental  Research Center, Analytical
     Quality Control Laboratory, Cincinnati,  Ohio 45268,  1973.

(8)  Webb,  R.G. and McCall, A. C.,  "Quantitative PCS Standards for Electron
     Capture Gas Chromatography."  Presented at  the 164th  National ACS
     Meeting, New York, August 29,  1972.   (Submitted to the Journal of
     Chromatographic Science for publication).

(9)  Goerlitz,  D.  F.  and Law,  L. M.,  "Note on Removal of  Sulfur Interferences
     from Sediment Extracts for Pesticide Analysis," Bulletin of  Environmental
     Contamination and Toxicology,  6,  9 (1971).

(10)  Mills, P.A.,  "Variation of Florisil  Activity:   Sample  Method for Measuring
     Adsorbent Capacity and its Use in Standardizing Florisil Columns,"
     Journal of the Association of  Official Analytical  Chemists.  51, 29 (1968).

-------
                           Table 1

                 Composition of Aroclor 1221

HPT8
11
14
16
19
21
28
32

P7
Uo
Total
Mean
Weight
Percent
31.8
19.3
10.1
2.8
20.8
5.4
1.4

1.7
93.3

Relative
6td. Dev.*>
IS. 8
9.1
9.7
9.7
9.3
13.9
30.1

48.8


Number of
Chlorines?
1
1
2
2
2
2] 65%
3J 15%
21 104
3J 90%
3
3

     •Retention time relative to p,p*-DDE«100.  Measured froc
first appearance of solvent.  Overlapping peaXs that axe
quantitated as one peak are bracketed.
     bstandard deviation of seventeen results as a percentage
of the mean of the results.
     CFroxn GC-MS data.  Peaks containing mixtures of isoroers
of different chlorine numbers are bracketed.

-------
                            Table 2

                 Composition of Aroclor 1232 C8>


HPT*
11
14
16
f20
Ul
28

32
37
40
47
54

58
70
78
Total
Mean
Weight
Percent
16.2
9.9
7.1
17.8
9.6

3.9
6.6
6.4
4.2
3.4

2.6
4.6
1.7
94.2

Relative
Std. Dev.b
3.4
2.5
6.6
2.4
3.4

4.7
2.5
2.7
4.1
3.4

3.7
3.1
7.5


Kunber of
Chlorinesc
1
1
2
2
2
21 40%
3J 60%
3
3
3
4
3] 331
4J 67%
4
4] 90%
5J 10%
4

     •Retention tiroe relative to p,pf-DD£«100.  Measured fror.
first appearance of solvent.  Overlapping peaXs that are
quantitated as one peak are bracketed.
     ^Standard deviation of four results as a Dean of the
results.
     CFron GC-MS data.  Peaks containing mixtures of isomers
of different chlorine numbers.are bracketed.

-------
                           Table 3

                 Composition of Aroclor 1242

HKT8
11
16
21
28
32
37
40
47
54
58
70

78
64
98
104
125

146

Total
Mean
Weight
Percent
1.1
2.9
11.3
11.0
6.1
11.5
11.1
8.8
6.8
5.f
10.. 3

3.6
2.7
1.5
2.3
1.6

1.0

98.5

Relative.
Std. Dev.b
35.7
4.2
3.0
5.0
4.7
5.7
6.2
4.3
2.9
3.3
2.B

4.2
9.7
9.4
16.4
20.4

19.9



Nuraber of
Chlorines0
1
2
2
21 25%
3J 75%
3
3
3
4
31 33%
4J 67%
4
41 so;
5J 10%
4
5
5
5
51 85%
6J 151
51 75%
6J 25%

     •Retention time relative to p,p*-DDE»100.   Measured  froa
first appearance of solvent.
     ^Standard deviation of six results as a percentage of
the »ean of the results.
     cFron CC-MS data.  PeaXs containing mixtares of iscsoers
Of different chlorine mu&ers are bracketed.
                                 (,0

-------
                            Table 4

                  Composition of Aroclor 1248


*KT*
21
26
32
47
40
47
54
58
70
78
84
96
104
112
125

146

Total
Mean
Weight
Percent
1.2
5.2
3.2
8.3
8.3
15.6
9.7
9.3
19.0
6.6
4.9
3.2
3.3
1.2
2.6

1.5

103.1

Relative
Std. Dev.b
23.9
3.3
3.6
3.6
3.9
1.1
6.0
5.6
1.4
2 • "*
2.6
3.2
3.6
6.6
5.9

10.0



Number of
Chlorines0
2
3
3
3
3*1 85%
4] 15%
4
31 10%
4J 90%
4
41 80%
5J 20%
4
5
5
41 10%
5J 90%
5
51 90%
6J 10%
51 85%
6J 15%

     •detention time relative to p,p'-DD£«100.  Measured from
first appearance of solvent.
     ^Standard deviation of six results as a percentage of
the mean of the results.
     cPron CC-MS data.  Peaks containing mixtures of isomers
of different chlorine numbers are bracketed.

-------
                            Table 5

                  Composition of Aroclor 1254


**!•*
47
54
58
70
84
98
104
125
14 £
160
174
203
232
Total
Mean
Weight
Percent
6.2
2.9
1.4
13.2
17.3
7.5
13.6
15.0
10.4
1.3
8.4
1.8
1.0
100.0

Relativew
Std. Dev.b
3.7
2.6
2.6
2.7
1.9
5.3
3.6
2.4
2.7
8.4
5.5
18.6
26.1


Number of
Chlorines
4
4
4
41 25%
5J 75%
5
5
5
5"! 70%
6J 30%
5T 301
6J 70%
6
6
€
7

     •Retention time relative to p,p'-DD£«100.  Measured froir
first appearance.of solvent.
     ^Standard deviation of six results' as a percentage of the
nean of the results.
     cFro» GC-MS data.  Pea3cs containing mixtures of isomers
are bracketed.

-------
                            Table 6

                 Composition of Aroclor 1260 Ct)


^&p^*fl
*\*%*
70
84
r 98
Ll 04

117
125
146
160
174
203

T232
[244

280
332
372
448
528
Total
Mean
Weight
Percent
2.1
4!7
3.8


3.3
12.3
14.1
4.9
12.4
9.3


9.8

11.0
4.2
4.0
.6
1.5
98.6

Relative
Std. Dev.b .
6.3
1.6
3.5


6.7
3.3
3.6
2.2
2.7
4.0


3.4

2.4
5.0
8.6
25.3
10.2


Number of
Chlorines0
5
5
-Id
5 60%
6J 40%
6
51 15%
6J 85%
6
6*1 50%
7J 50%
6
61 10*
7j 90%
1«
6 10%
7J 90%
7
7
8
8
8

     •Retention tine relative to p,p'-DDE«100.  Measured froE
first appearance of solvent.  Overlapping peaks that are
quantitated as one.peak arc bracketed.
    .•^StAndard deviation of six results as a mean of the
re suits .-
     ^From CC-MS data.  PeaXs containing mixtures of isoners
Of different chlorine nxunb&rs are bracketed.         *
     ^Composition determined at the center of peak 104.
     ^Composition determined at the center of peak 232.

-------
COMPRESSED
AIR	
SUPPLY
                                            PRESSURE
                                            'GAUGE
               SHUT-OFF
                VALVE
REGULATOR
NEEDLE o
VALVE w
                               I em
         FLEXIBLE
          TUBING
               SILICA GEL
                  5 err
                     I em
                                   I  10/30
                                .  15ml
                                 RESERVOIR
                                   §  10/50
                                   23 cm  i  4.2 mm t.D.
                                   2 cm  • 2 mm I.D.
           FIGURE  I.   WICROCOLUMN  SYSTEM
                         X-H

-------
 o
 o
r


c
a.
*j
c
                                      C\J
                        1V101  JO  lN3DH3d


                              I- c^S"

-------
               37
                  AROCLOR 1242
Pi£urc  3. Column: 3% OV-1,  Carrier Gas: Nitrogen at  60  ml/min,
         Column  Temperature: 170 C, Detector: Electron  Capture

-------
 \
                              AROCIOR 1254
                             125
                                                      JJ2
Figure  4. Column: 3% OY-1, Carrier Gas: Nitrtgen it SO nl/nin,
         Column Temperature:  170 C, Detector: Electron Capture.

-------
                               AROCLOR 1260
Fifure  5. Column: 3% OY-1. Carrier fits: Nitrogen it SO ul/nin,
         Column Temperature: 170 C, Detector: Electron Capture

-------
                      AROCLOR 1242
              j
I
                                   I
1
I
I
i
                           24
              i      I      12      15      II     21
                     IEUKTIOK TIKE  IK MIKCTES
      S. Ctluno: 1.5% DM7 * M5% |F-1, Cirrfir gts:
it SO • I/lie,  Column Timpintin: 200 C, Pitictir: Eij:trn (iptire,

-------
  e
  r
  ^ S
. if
  55
  w

- i

-------
 *
 1
 r»

ii
I *
 I
 i

-------
           } tRT it first M«fc«<7?  J

           TES/          \«D
    Is tken i fistiict
    ink with IRT 71?
                  III 47-51?
     TES
/
TES
                          X
                           *
  Use 1242 fir
          14
        Use 1242  fir
       peikjl IRT 70
      \       I/
   Use 1254
   fir fills
   1 IRT 104
         Is then i listiict
         ftik witk IRT 117?
                        •se 1260 fir
                         all piiks
       TES
           RO
                lite 12S4 fir ill
                    l IRT 174
    • si 1260 fir
    all itker pnks
Fi£urt 9. ChromatO£rim Division Flowchart [8].

-------
Procedure Code: 14

Variations:  1-11

Title:  Pentane Extraction of Organics in Wastewaters for GC Analysis

1.   Summary

     The following procedure is adapted from that reported by Henderson e_t_
     ajUl).  It has been successfully applied to the analysis of benzene,
     toluene, ethylbenzene, and halogenated volatile compounds, including
     chloroalkyl ethers, in a wide variety of industrial wastewaters.  A snail
     volume of water sample is shaken with pentane in a headspace-free bottle.
     The pentane layer is then injected, without concentration, into a gas
     chromatograph equipped with electron-capture or Hall detector for halogen-
     ated compounds, or with an FID detector for detection of non-halogenated
     compounds.  This method differs from the microextraction procedure (Pro-
     cedure Code #7) in the volume of extracting solvent and to sample ratio;
     this ratio is larger in Procedure #14 than in #7.  Alternative procedures
     for many of the compounds analyzed for by procedures #14 and #7 are given
     in the purge and trap method Procedure Code #8.

2.   Sample Collection

     Samples are collected in 40-ral vials, which are filled to overflowing,
     then capped with a teflon-coated rubber septum lined with aluminum foil.
     The septum is held in place by a screw-cap retainer with a center hole.
     The samples are held at 4 C until analysis.

3.   Interferences

     3.1  Solvents, reagents, glassware, and other sample-processing hardware
          may yield discrete artifacts and/or elevated baselines, which cause
          misinterpretation of gas chromatograms.   All these materials must be
          shown to be free of interferences under the conditions of the analysis,
          by running method blanks with each set of samples.

     3.2  Because of the non-specificity of the extraction procedure, co-
          extracted compounds other than those to be analyzed may interfere.
          The use of GC/MS or other confirmatory techniques, such as second GC
          columns, multiple-phase equilibration (2), or solvent-partition tests
          (3), may be required to confirm the qualitative analysis.

4.   Reagents, Solvents, and Standards

     4.1  Sodium sulfate - (ACS) Granular, anhydrous (conditioned at 400 C for
          4 hours)

     4.2  Pentane-nanograde; redistill in glass if necessary.

     4.3  Organic compounds standar        oercent purity).

-------
5.   Sample Preparation

     5.1  Allow the sample to come to room temperature.  (If compositing is
          required, pour the required volumes together in a chilled (ice-bath)
          flask while the samples are still cold from storage; then place the
          composite in a vial, seal the vial, and warm to room temperature).

     5.2  Add 5.0 ml of sodium sulfate saturated organic-free water directly to
          the vial by piercing the septum with two 10-ml syringes (22-gauge
          needles).  One syringe should contain the water solution, and the
          other should be empty.  Insert the vial, inject the sodium sulfate
          solution, and discard the displaced water.

     5.3  Add 5.0 ml of nanograde pentane to the vial through the sane syringe
          system described in 5.2.

     5.4  Shake the vial for 15 minutes at 500 rpm with a gyratory platform
          shaker or equivalent.

     5.5  Withdraw portions of the pentane layer as required for GC analysis.
          (See method variations in Section 8).

6.   Quality Control

     6.1  Duplicate and spiked sample analyses are recommended as quality
          control checks.

     6.2  With each set of samples, extract and analyze a method blank, using
          organic-free water.

7.   Calculation of Results

     7.1  Determine concentrations by using the absolute calibration procedure
          or by using a suitable internal standard calibration procedure.
          Henderson e£ a_l (1) found that an internal standard could be conven-
          iently added to the pentane used for extracting samples and standards.

                                (A) (B) (Vt)
          micrograms/liter  =  TV~)—(V~~)	


           ,          .   ng standard
          where:    A == —*—i—-.	
                        standard area

                    B = Sample aliquot area

                    V. « Volume of extract injected  (ul)

                    V  » Volume of total extract  (yl)

                    V  = Volume of water extracted  (ml)

-------
     7.2  Report results in raicrograms per  liter without correction for recovery
          data.   When duplicate  and  spiked  samples are analyzed, all data
          obtained .must be reported  on Analytical Quality Control Report sheets.

8.   Variations  to Procedure Code  //14:
     Variation No.
Effective
  Date

June 1977
                         June 1977
                         June 1977
     Description

GC Column:  3% OF-1 on 80/100"mesh
Chromosorb WHP in a 6' x 2 mra ID glass
column.

Temperature:  35°C isothermal

Injector:  200°C

Detector:  Ni63 at 300°C

Carrier:  Argon/methane at 40 ml/min.

(Source:  EEI)

GC Column:  20% OV-101 +0.1% Carbowax
1500 on 80/100 mesh Chromosorb W in a
6* x 2mm ID glass column.

Temperature:  65 C for 30 min; 65 C to
110 C at 10°/min.; hold at 110 C for
60 min. if dichlorobenzene is present.
                                                       at  325°C
               Injector:  175 C

               Detector:  Ni

               Carrier:  Argon/methane at 40 ml/min.

               (Source:  EEI)

               GC Column:  20% OV-101 + 0.1% Carbowax
               1500 on 80/100 mesh Chromosorb W in a
               6' x 2 mm ID glass column.

               Temperature:  65 C isothermal

               Injector:  175°C

               Detector:  Ni63 at 300°C

               Carrier:  Argon/methane at 40 ml/min.

               (Source:  EEI)
                                t-ts

-------
June 1977      GC Column:  20% OV-101 + 0.1% Carbowax
               1500 on 80/100 mesh Chromosorb W in a
               6* x 2 mm ID glass column.

               Temperature:  65°C to 110°C at 10°C/min.

               Injector:  175°C

               Detector:  Ni63 at 325°C

               Carrier:  Argon/methane at AO ml/min.

               (Source:  EEI)

June 1977      GC Column:  20% OV-101 +0.1% Carbowax
               1500 on 80/100 Chromosorb W in a 6' x 2
               mm ID glass column.

               Temperature:  70 C for 16 min., then
               10°C/min to 110°C; hold 15 min.

               Injector:  150°C

               Detector:  Ni63 at 325°C

               Carrier:  Argon/methane at AO ml/nin.

               (Source:  EEI)

May 1978       GC Column:  20% OV-101 + 0.1% Carbowax
               1500 on 80/100 Chromosorb W in a 6' x 2
               mm ID glass column.

               Temperature:  90 C isothermal

               Injector:  175°C

               Detector:  Ni63 at 300°C

               Carrier:  He at 50 ml/min

               (Source:  EEI)

June 1977      GC Column:  20% SF-96 on 60/80 mesh
               Chromosorb P-AW on a 6' x 1/8" OD
               stainless steel column preceded by a 6'
               x 1/8" OD stainless steel column packed
               with Chromosorb 102.

               Temperature:  75 C for 5 min.; program
               to 150°C at A°C/min.
 J

-------
10
11
               Injector:  200 C

               Detector:  FID at 250°C

               Carrier:  He at 35 ml/min.

               (Source:  EE1)

June 1977      GC Column:  10% TCEP on 60/80 mesh
               Chromosorb P in a 6' x 1/8" OD stainless
               steel column.

               Temperature:  90  isothermal

               Injector:  145°C

               Detector:  FID at 190°C

               Carrier:  Ke at 35 ml/nin.

               (Source:  EEI)

June 1977      Same as  (8), except Temperature = 70 C
               isothermal

               (Source:  EEI)

May 1978       GC Column:  10% TCEP on 100/120 mesh
               Chromosorb P-AW in an 8' x 1/8" OD
               stainless steel column.

               Temperature:  70 C isothermal

               Injector:  175°C

               Detector:  FID

               Carrier:  He at 30 ml/min.

Jan. 1979      GC Column:  10% TCEP on 100/120 Chromo-
               sorb P-AW in an 8' x 1/8" OD stainless
               steel column.

               Temperature:  75 C isothermal

               Injector:  190°C

               Detector:  FID at 190°C

               Carrier:  He at 30 ml/min.

               (Source:  EEI)

-------
9.   References

     1.   Henderson, J.E., Peyton, G.P., and Glaze W.H., "A Convenient Liquid-
          Liquid Extraction Method for the Determination of Haloraethanes in
          Water at the Parts-Per-Billion Level", in:  Identification and Analysis
          of Organic Pollutants in Water. Ed. L.H. Kieth, p 105, Ann Arbor
          Science, Ann Arbor, MI  (1977).

     2.   Beroza, M., and Bowman, M.C., "Identification of Pesticides at Nanogram
          Level by Extraction p-Values", Anal. Chem.. .37, 291-292 (1965).

     3.   McAullife, C., "GC Determination of Solutes by Multiple Phase Equili-
          bration", Chem. Tech., Jan., 46-51 (1971).

-------
Procedure Code: 14                                                     ? / / 7 / 7 9
     Variation 112 (Source: EEI) May, 1979
     1.  Allow the sample * to cone to room temperature,  (If compositing
         1> required pour the required volumes together in a chilled (ice-
         bath) flask while the samples are still cold from storage; then
         place the composite in a vial, seal the vial, and warm to room
         temperature^-
         *  40 ml vial or 125 ml vial, as required.
     2.  Remove 4.0 ml of sample with a 50-ml syringe.  Then add 3.0 ml of
         •odium sulfate saturated organic-free water using another 10 ml
         syringe.  This creates a 1.0 ml headspace in the vial.
     3.  Add 2.0 ml of nonograde pentane to the vial by piercing the septum
         with two 10-ml syringes.  One syringe should contain the pentane,
         and the other should be empty.  Insert the vial, inject the pentane,
         and discard the displaced water.
     4.  Shake the vial for 15 minutes at 500 rpm with a gyratory platform
         •baker or equivalent.
     5.  Withdraw portions of the pentane layer as required for GC analysis
     6.  GC Conditions (Non-Balogenated Compounds)
           GC Column:                10Z 1, 2, 3-tricyanoethoxypropane
                                     (TCEP) on 100/120 chromosorb PAW,
                                      8* x 1/8" OD stainless steel column
         Temperature:                75°C isothermal
         Injector   :                190°C
         Detector   :                FID at 300°C
         Carrier    :                H2 at 30 ml/mln
         GC Conditions (Halogenated Compounds)
         GC Column  :                20Z OV-101 + 0,11 Carbowax 1500
                                     on 80/100 Chromosorb VAW  6' x 2 mm
                                     glass
         Temperature:                60°C for 10 min.  to 130°C at 15°/min
         Injector   s                130°C
         Peteetor   i                Hi63 at 270°C
         Carrier    t                N2 at 30 mil/mln
     Variation 113
          Same as Variation 112, except:
            GC Temperature:  60 C Isothermal for halogenater compounds on 20Z
                     Q      OV-101 + 0.1Z Carbowax 1500 Column.

-------
Procedure Code:  15

Title:    Acid Extraction Procedures for Phenols

Variations:    1-2

1.   Summary
          The methods described here are similar to the EPA acid-extraction
     procedure used during the screening of industrial effluents (1).  Two
     variations are presented.  In Variation #1, the sample is extracted at
     high pH (above pH 11) to remove base-neutral compounds, then is re-extracted
     at pH<2.  This extract can be analyzed for phenols with GC/FID, or it can
     be back extracted into sodium hydroxide solution for cleanup by an A-26
     anion exchange resin procedure similar to that of Chriswell et^ ai_. (2).
     (See Procedure Code #5.)  The final methylene chloride or hexane extract
     is analyzed by GC/FID.  In Variation #2, the sample is acidified and
     extracted, as in the screening protocol (1), with methylene chloride.
     GC/FID is then used instead of GC/MS for compound detection.

          The following table gives the proper code numbers for each variation:

     Variation      Effective
        No.            Date             Description
        1           February 1979       Southwest Research
                                        Method with A-26 resin

        2           June 1977           Envirodyne/EPA Screening Method

     2.0  Variation #1   (Southwest Research Method):

     2.1  Measure a 1-liter water sample into a 1-liter graduated cylinder,
          and pour into a 2-liter separatory funnel.

     2.2  Adjust the pH to 11 or greater with 6N NaOH.  Use multi-range pH
          paper for measurement.

     2.3  Extract 3 times with 60 ml of methylene chloride.  (This extract
          may be required for PAH (Procedure Code #17) or Pesticides-Phthalates
          (Procedure Code #3); if so, do not discard.)

     2.4  Adjust the pH of the water sample to 2 or less with 6N HC1, using
          pH paper as indicator.

     2.5  Extract 3 times with 60 ml of methylene chloride each time.

     2.6  Combine the extracts in a 500-ml Kuderna-Danish evaporator, and
          concentrate to about 5 ml on a steam bath.

-------
2.7  Solvent exchange to hexane (using 100 ml of hexane), and reconcentrate
     to about 2 ml, using a stream of dry nitrogen for the final step.

2.8  Cleanup with A-26 Resin:

     Since some neutral compounds may still be present in the hexane
     extract and could cause artifact peaks during gas chromatography, the
     following steps can be taken to remove neutral organics from the
     phenols:

     2.8.1     Pour the hexane extract into a 30-ml separatory funnel.

     2.8.2     Extract 5 times with 5 ml of 0.IN NaOH, and combine the
               NaOH extracts.

     2.8.3     Prepare an Amberlyst A-26 column (10 on x 0.6 cm) from
               resin that has been cleaned and sieved (2).  Place the
               column in the hydroxide form by washing the column with 20
               ml of 0.1N NaOH, then with 20 ml of organic-free water.
               Check the first portion of water with pH paper (pH should
               be <9) to ensure that the resin is saturated with the base.

     2.8.4     Flow the combined 0.1N NaOH extracts into the resin, down
               to the top of the column.

     2.8.5     Add 25 ml basic methanol (2 ml of 2N NaOH in 23 ml of
               methanol), then 25 ml of organic-free water; discard the
               eluate.

     2.8.6     Place a 125-ml separatory funnel under the column, and
               elute with 25 ml of 4N HC1 followed by 25 ml of organic-
               free water.

     2.8.7     Extract the eluate from step 2.8.6 with 25 ml of methylene
               chloride, and drain the methylene chloride into a second
               125-ml separatory funnel.  Save the aqueous layer.

     2.8.8     Elute the resin with 25 ml of acetone:water (5:1, v:v),
               followed by elution with 25 ml of organic-free water,
               into the second separatory funnel with the methylene
               chloride.

     2.8.9     Extract by shaking for two minutes, and drain the methylene
               chloride layer into a 500 ml K-D evaporator.  Save the
               aqueous layer.
                              T -

-------
          2.8.10    Repeat  steps  2.8.7  and  2.8.9 twice more,  with fresh 25-ml
                    portions  of methylene chloride.*

          2.8.11    Concentrate the  combined  organic extracts to about 5 ml
                    on a steam bath.

          2.8.12    Solvent exchange the methylene chloride extract into
                    hexane  (using 100 ml of hexane), and concentrate to 1 ml
                    for analysis, using a stream of  dry nitrogen.

     2.9  GC Conditions:

          Column:    1% SP-1240DA  on  100/120 mesh Supelcoport in a 6' x 2mm
                    ID glass  column.

          Temperature:   80°C for 2  min, 80°  to 180° at 8°/min, hold at
                         180° for 15 min.

*This step differs from the Chriswell et al.   (2) procedure,  which does not
 use the two additional liquid-liquid extractions of the aqueous layers.

          Injector:     250°C

          Detector:     FID at 250°C

          Carrier:     He at 60 ml/min.

     2.10 Quality Control and Quantification:

          2.10.1    Duplicate and spiked sample analyses are recommended as
                    quality controls.  Each set of sample analyses is preceded
                    by a method  blank using organic-free water.

          2.10.2    Calculations  of phenol  concentration are determined by
                    using the following calibration procedure, or by using
                    another suitable relative calibration procedure.
                                          (A) (B) (Vt)
                           Micrograms/1 ~
                                           (V±) (Vs)
                            yg Standard
                            Standard area
                     B   =  Sample aliquot area

                     V   -  Volume of extract injected (yl)

                     V   -  Volume of total extract (yl)

                     V   «  Volume cf water extracted (ml)
                                 x-

-------
3.0  Variation #2  (Envirodyne/EPA Screening Method).

     3.1  Transfer a 1000 ml sample to a separatory funnel, and adjust the
          pH to 2 or less with 6N HC1.

     3.2  Extract serially with 100 ml, 50 ml, and 50 ml of distilled-in-
          glass methylene chloride.

     3.3  Dry and filter the solvent extract by passing it through a short
          column of sodium sulfate.  (The sodium sulfate should be prewashed
          in the column, using methylene chloride.)  If necessary, the
          sodium sulfate should be pretreated by heating at 500 C for 2
          hours in an oven to remove interfering organics.

     3.4  After passing the extract through the sodium sulfate, rinse the
          column with methylene chloride and add the rinsings to the
          extract.

     3.5  Concentrate the extract to 5-10 ml in a 500 ml Kuderma-Danish
          apparatus with a 3-ball macro Snyder column and 10 ml receiver
          tube attached.  Cool to room temperature, add fresh boiling
          chips, and attach a micro Snyder column to the receiver tube.
          Evaporate carefully to 1.0 ml or to the point when active
          distillation ceases.

     3.6  GC Conditions:

          Column:   3% OV-17 on 80/100 mesh Chromosorb in a 6' x 1/8"
                    OD stainless steel column.

          Temperature:   160° to 250°C at 8°/min.

          Injector:      300°C

          Detector:      FID

          Carrier:       He at 35 ml/min.

     3.7  Quantification is achieved by comparing peak areas of the sample
          extracts to the peak area of a standard on the chromatograph.
          Sample spikes are carried through the procedure to determine the
          recovery of the various phenols.

-------
4.0  References
          1.    Sampling and Analysis Procedures for Screening of industrial
                Effluents for Priority Pollutants, U.S. EPA Environmental
                Monitoring and Support Laboratory, Cincinnati, OH, March
                1977 Revised April, 1977.

          2.    Chriswell, C.D., Chang, R. C., and Fritz, J. S., "Chromato-
                graphic Determination of Phenols in Water," Anal. Chem.,
                47 1325-1329 (1975).

-------
Procedure Code:  16

Title:  Analysis of Nitroaromatics

1.   Summary

          The method recommended for Nitroaromatics is that proposed by the
     U.S. EPA, Method 609 (1).  A 1-liter sample is extracted with methylene
     chloride.  The extract is dried and exchanged to toluene while being
     concentrated to 1.0 ml.  Nitrobenzene is measured by FID/GC; dinitro-
     toluenes are measured by EC/GC.  A florisil cleanup is provided to
     eliminate interferences.

2.   References

     1.   Methods for Organic Compounds in Municipal and Industrial Wastewater^
          U.S. EPA Environmental Monitoring and Support Laboratory, Cincinnati,
          Ohio, March (1979).

-------
Procedure Code:   17

Variations:   1-5
REVISED
Title:   Analysis  of Polynuclear Aromatic Hydrocarbons  (PAH) in Industrial
        Wastevater

1.   Summary

     Four procedures for analysis of PAH compounds  in  industrial vastewaters,
     •re described in this Procedure Code.   In Variations 1 and 2, hexane is
     used as an extraction solvent, followed by silica-gel cleanup.  In
     Variation 3, methylene chloride is used in a continuous extractor,
     followed by  a DHSO cleanup.  Variation 4 also  involves methylene chloride
     extraction, with a separately funnel,  followed by Florisil cleanup.  The
     following table lists the method variations:

     Variation No.       Effective Date          Description
          1             June,    1977           EEI  Procedure a

          2             January, 1979           EEI  Procedure b

          3             January, 1978           MRI  Procedure
          4             March,   1979           SwRI Procedure
          5             May,     1979           EEI  Procedure c

     The SwRI extraction/cleanup procedure  (Variation  14) is similar to the
     extraction/cleanup procedure (Method 610) proposed by the U.S. EPA (1),
     except for the use of a Florisil cleanup by SwRI  instead of the silica-gel
     cleanup in the EPA method.

     Since the separation of some pairs of  PAH compounds by conventional phase
     packed-column gas chromatography is not possible, several options are
     available to the analyst.  With liquid chromatography and DV or fluorescence
     detection (EPA Method 610) , all the priority pollutant PAH compounds can
     be separated.  Capillary-column GC has also been  shown to separate them.
     Liquid-crystal, packed-column GC has been shown to separate 2-ring to 6-
     ring PAH compounds more effectively than conventional phase packed columns.
     The difficult separations  involving priority pollutants are the following
     pairs:
          a.   benzo(a) pyrene  from perylene
          b.   anthracene from  phenanthrene
          c.   chrysene from benzo(a) anthracene
          d.   benzo(b) fluoranthene from benzo(k)  fluoranthene

     However, there may be other PAH compounds that interfere with the priority
     pollutant PAH compounds in any of the  analytical  options presented here.

2.   Variation tfl (EEI Procedure a)

     2.1  Extract 500 ml of sample with one 100-ml  portion and two 50-ml portions
          of nanograde hexane.  Concentrate the combined extracts to 1 ml using
          a Kuderna-Danish evaporator followed by a stream of nitrogen.

-------
     2.2  Transfer the extract to the top of a silica-gel column prepared in
          a 4 1/2" "champagne" column (30 ml reservoir; Supelco> Inc.) with
          about 4" of 60/80 mesh Davison Grade-12 silica gel.  The silica gel
          is activated by heating at 100 C for 20-30 minutes before use.

     2.3  Elute the column with enough hexane to collect 5 ml of eluate; save
          this eluate as Fraction #1.  Elute with benzene and collect two 2-ml
          fractions (#2 and #3).  Finally, elute with enough benzene to collect
          a 4-ml Fraction (#4).

     2.4  Analyze benzene fractions #2 and #3 for PAH compounds by GC._  If
          high concentrations are found in these fractions, analyze the last
          benzene fraction (#4).  Compute the volumes and concentrations in
          each fraction, to determine the total weight of PAH recovered; then
          compute the apparent concentration in the original sample, uncor-
          rected for recovery.  Report the results of duplicate and spiked
          samples on quality-control sheets.

     2.5  GC Conditions:

               GC Column:    3Z OV-17 on 80/100 mesh Chromosorb W-AW in a 6'
                             x 1/8" OD stainless steel column.

               Temperature:  140 C for 10 min., then programmed at 8 /min.
                             for 15 minutes, followed by 4°/min. to 280°C.

               Injector:     280°C

               Detector:     FID at 300°C

               Carrier:      He at 50 ml/min.

               Note:  The four pairs of PAH compounds listed in Section 1
                      cannot be separated on this column.

3.   Variation #2  (EEI Procedure b)

     3.1  Extract 500 ml of sample with one 100-ml portion and two 50-ml
          portions of nanograde hexane.  Concentrate the combined extracts to
          1 ml using a Kuderna-Danish evaporator followed by a stream of
          nitrogen.

     3.2  Transfer the extract to the top of a silica-gel column prepared in a
          4 1/2" "Champagne" column (30 ml reservoir; Supelco, Inc.) with
          about 4" of 60/80 mesh Davison Grade-12 silica gel.  The silica gel
          is activated by heating at 100°C for 20-30 minutes before use.

     3.3  Elute the column with enough hexane to collect a 3-ml fraction (#1).
          Next, collect a 2-ml fraction of hexane eluate (Fraction #2).   Elute
          with benzene and collect 2 ml of eluate; combine this 2-ml benzene
          fraction with hexane Fraction #2.  Continue to collect benzene
          eluate until 7 ml of additional benzene eluate have been collected
          in a separate fraction (Fraction

-------
     3.4  Analyze Fraction #2 for an estimate of the quantity of PAH compounds
          present in it.   If  the indicated amount is high, it may be necessary
          to analyze Fraction $1 also.   Otherwise, discard Fraction //I, and
          analyze Fractions #2 and #3 as noted in 3.5.

     3.5  Gas Chromatography  Conditions

          a.   for Fraction #2 (and til if needed):

               GC Column  :     3% OV-17 on 80/100 Chromosorb W-AW in a 6' x 1/8"
                              stainless steel column.

               Temperature:   Hold at 110°C for 12 min., increase to 240°C at
                              10°/min., hold at 240°C for 5 min., and cool 5
                              minutes.

               Injector:       280°C

               Detector:       FID at 300°C

               Carrier:        He at 35 ml/min.

          b.   for Fraction #3

               GC Column:      2.5% BMBT liquid crystal on 100/120 mesh Chromosorb
                              WHP in a 4' x 1/8" stainless steel column (2).

               Temperature:   185°C for 2 min., increase to 240°C at 4°/min.,
                              and hold for 10 minutes.

               Injector:       265°C

               Detector:       FID at 265°C

               Carrier:        He at at 40 ml/minute.

4.   Variation #3  (MRI  Method) (3)

          4.1  Measure and record the volume of sample.  Adjust the pH to 11.0
               or greater with 5N. NaOH.  (If spiking is to be done, add an
               aliquot of mixed PAH standard in acetone before pH adjustment).

     4.2  Transfer the sample to a specially-designed liquid-liquid extractor
          system (see Figure  1), which contains methylene chloride (Burdick and
          Jackson, Muskegon,  Michigan).  Extract overnight (16-18 hours).
          After extraction, drain off the methylene chloride and pass it
          through a drying tube containing anhydrous sodium sulfate (ACS
          grade, dried at 600°C and stored at 110°C) into a Kuderna-Danish (K-
          D) evaporator with  10 ml of hexane in the receiver.  Add boiling
          chips and concentrate to about 10 ml.

-------
     A.3  Transfer the hexane to a 125-ml separatory funnel, and extract
          sequentially three times with 10 ml of dimethylsulfoxide (DMSO).
          Combine the DMSO extracts, add 60 ml of deionized distilled water,
          and extract three times sequentially with 90 ml of hexane in a 250-ml
          separatory funnel.

     4.4  Wash the hexane layer once, with an equal volume of water,  to insure
          removal of the DMSO; then pass the hexane solution of PAH compounds
          through a drying tube of sodium sulfate into a K-D evaporator, and
          concentrate to less than 5 ml.  Further concentrate to 1.0 ml under a
          gentle stream of prepurified nitrogen.  This cleanup procedure is as
          described in Natusch and Tomkins (3).

     4.5  Analyze the extract of PAH compounds on a 30-m wall-coated open
          tubular capillary column (WCOT) as follows:

               GC Column:     SE-30 WCOT capillary column, 30 in. long
                              (min: 2,500 effective plates/m) (Supelco, Inc.).

               Temperature:   50°C for 1 min., then 8°C/min. to 260°C; hold
                              12 min. at 260°C.

               Injector:      180°C

               Detector:      250°C, FID

               Inlet Split:   120:1 ratio

               Carrier:       1.0 ml/min. N-; detector makeup flow 35 ml
                              min. N_.

          Figure 2 shows a chromatogram from a 20 mg/1 standard (4 yl) injection.

5.0  Variation 14  (SwRI Method)

     5.1  For extraction of PAH compounds, the EPA (1) procedure of Method is
          used, as follows:

          a.   Measure a 1-liter water sample into a 1-liter graduated cylinder.

          b.   Pour into a 2-liter separatory funnel.

          c.   Adjust the pH to 11 or greater with 6N NaOH, using multirange pH
               paper for measurement.

          d.   Extract 3 times with 60 ml of methylene chloride.

          e.   Combine extracts in a 500-ml Kuderna-Danish evaporator.

          f.   Evaporate to about 5 ml on a steam bath.

          g.   Solvent exchange to hexane, using 100 ml of hexane.

          h.   Concentrate to about 2 ml under a stream of dry nitrogen.
                                        n

-------
5.2  Florisil preparation:   Heat  Florisil to 400 C overnight; add 3% HJO
     (m/m) and .mix thoroughly;  let stand for at least 2 hours.  Use a
     10mm ID column chromatography tube with a glass frit at the bottom
     and a teflon stopcock  (Kontes Catalog K420540-0213).

5.3  Cleanup of  the hexane  extract is performed as follows:

     a.   Add lOg of Florisil  (32 H.O) to the column chromatography tube,
          and tap to settle.

     b.   Add about 1cm of  anhydrous sodium sulfate to the top of the
          column.

     c.   Add hexane to the column and pre-elute 40 ml; discard the
          eluate.

     d.   Drain  until the solvent level is even with the top of the
          sodium sulfate.

     e.   Add sample (in about  2  ml of hexane) to the top of the column.

     f.   Drain  the sample  into the column to the top of the sodium
          sulfate.

     g.   Wash the sides of the tube with about 2 ml of hexane, and drain
          to the top of the sodium sulfate.

     h.   Place  a 500-ml Kuderna-Danish evaporator under the column.

     i.   Collect 20 ml of  hexane eluate in th* K-D evaporator.

     j.   Evaporate the hexane  eluate to less than 10 ml on a steam bath.

     k.   Transfer the concentrate to a solvent-rinsed 4-dram vial with a
          teflon cap liner.

     1.   Adjust the final  volume of concentrate to 5 ml.  Evaporate
          under  nitrogen if necessary.  (This is Fraction I, and contains
          naphthalene).

     m.   Place  another 500 ml  K-D evaporator below the column.

     n.   Collect 40 ml of  hexane eluate in the K-D evaporator.

     o.   Evaporate to less than 10 ml on a steam bath.

     p.   Transfer the concentrate to a solvent-rinsed 4-dram vial with a
          teflon cap liner.

     q.   Adjust the final  volume of concentrate to 5 ml.  Evaporate
          under nitrogen if necessary.  (This is Fraction II, and contains
          the other PAH compounds).
                               x-qc

-------
5.4  Gas  chromatography  is performed  as  follows  for  Fraction 1:

          GC Column:          10.6 m  x 0.25 mm ID  SP2100  Wall-Coaced open
                              tubular capillary  column.

          Inlet Split Ratio:  100:1

                              300°C,  FID
Detector:

Injector:

Carrier:

Temperature:

     (used for naphthalene)
                              250°C
                              He at 0.75 ml/min. after split

                              95°C for 8 min., then 32°/min.  to 270°C;
                              hold 15 min. at 270°C.
5.5  Gas chromatography for Fraction 2 is performed in the same manner as
     for Fraction 1, with only a change in the temperature program:
          Temperature:   110°C to 270 C at 4°/min.; hold at 270°C for
                         15 minutes,  (used for fluoranthene, acenaphthene,
                         anthracene, fluorene, phenanthrene, pyrene.)

5.6  Gas chromatography for 1,2-Benzanthracene, 3,4-Benzopyrene,
     3,4-Benzofluoranthene, 11,12-Benzofluoranthene, Chrysene, 1,2-
     Benzoperylene, Indeno-(l,2,3-c,d) - pyrene:

          GC Column:  1.5% SP301
                      (BPhBT liquid crystal)  (A) on 100/120 mesh
                       Supelcoport in a 6' x  2 mm ID glass column
Temperature:

Detector:

Injector:

Carrier:
                        265 C Isothermal

                        300°C
                        300°C
                        He at 30 ml/min
Variation #5  (EEI Procedure c)

6.1  Extract 2 liters of sample with one 100-ml and two 50-ml portions of
     nanograde hexane.  Dry the combined extracts over sodium sulfate,
     and concentrate to 2 ml using a Kuderna-Danish evaporator, followed
     by a nitrogen stream.

6.2  Split the extract into two one-mi portions.  Analyze one portion-at
     the following GC conditions (used for naphthalene and acenanhthylene):

-------
               GC Column:     3% OV-17 on 80/100 mesh Chromosorb WHP in a
                             6' x 1/8" OD stainless steel column.

               Temperature:   Hold at 80"C for 7 min.; increase to 120°C at
                             4°/min.; hold for 5 min. cool 5 min.

               Injector:      185°C

               Detector:      FID at 300°C

               Carrier:       He at at 40 ml/min.

     6.3  Transfer the other one-mi portion of hexane extract to the top of a
          silica gel column  prepared in a 4%".  Champagne column (30-ml reservoir,
          Supelco, Inc.) with about 4 in. of 60/80 mesh, Davison Grade-12
          silica gel.   (The  silica gel is activated by heating at 100°C for
          20-30 minutes before use.)

     6.4  Elute the column with 5 ml of hexane, and discard the eluate.  Then
          elute with 50 ml of 40 percent methylene chloride in hexane.
          Concentrate this eluate to 1 ml for GC analysis at the following
          conditions (used for anthracene and phenanthrene):
               GC Column:



               Temperature:

               Injector:

               Detector:

               Carrier:
2.5% BMBT liquid crystal on 100/120 mesh Chromosorb
WHP in a -4' x 1/8" O.D. stainless steel column
with a 7" precolumn of 2.5% BMBT

185°C isothermal

265°C

FID at 300°C

He at 40 ml/min.
7.   Quality Control
     Duplicate and spiked sample analyses are recommended as quality control.
     Each set of sample analyses is preceded by a method blank using organic
     free water.

-------
8.   Calculations

     The concentration will be determined by using the absolute calibration
     procedure described below or a suitable relative calibration procedure.

                     (A) (B) (V2)

           8/1    "   0^) (Vg)


                  .   ng standard
                     standard area

          B       «  Sample aliquot area

          V.      •  Volume of extract injected ( 1)

          Vt      -  Volume of total extract ( 1)

          V       »  Volume of water extracted (ml)


9.   References

     1.   Methods for Organic Compounds in Municipal and Industrial Wastewater.
          U.S. EPA Environmental Monitoring and Support Laboratory, Cincinnati,
          Ohio, March, 1979.

     2.   Janini, G.M. Johnston, K., and Zielinski, W.L., "Use of a Nematic
          Liquid Crystal for Gas-Liquid Chromatographic Separation of Polyaromatic
          Hydrocarbons", Anal.  Chem., 47_, 670-674 (1975).

     3.   Natusch, D.F.S.,  and Tonkins, B.A.,  Isolation of Polycyclic Organic
          Compounds by Solvent Extraction with Dimethyl Sulfoxide", Anal.
          Chem.. .50, 1429-1434 (1978).

     4.   Janini, G.M.,  Muschik, G.M., Schroer, J.A.,  and Zielinski, W.L., Jr.
          "Gas-Liquid Chromatographic Evaluation and Gas-Chromatography/Mass
          Spectrometric Application of New High-Temperature Liquid Crystal
          Stationary Phases For Polycyclic Aromatic Hydrocarbon Separations,"
          Anal. Chem.. .48,  1879-1883 (1976).
                                             3

-------
                   6mm
                   Glom  Red
TFE  Reducing Union,
10-6mm ftomd  Through et 6mm
                 25mm I.D.
               Glass Drip Tip      •
       TFE Reducing Union, 10-omm,
        lonxf Through at 6mm
                    $24/40	».H
                  250ml
                                                                  1*5/50
 TFE  Stopcock


 TFE  Reducing  Union,  10-emnt



TFE Tubing, 6mm O.D.
                                        Figure 1.   MRI
                           Continuous  Liquid-Liquid Extractor
                             for  Use with Methylene Chloride

-------
           «H
     3
     00
     •


     CO
>    S1
  *-. *o
i   S
0

s
      pt
      o

     00
      M


      I
                                                            Acenaphthcne
                 Acenaphthylene
Phenanthrene

  Anthracene
                                      Fluoranthene


                                   Pyrene
                    Chryaene


                   Benzo(a)anthracene
                                       I-95-

-------
                                  Chapter  II


     Quality Assurance Program for the Organic Chemicals Industry
                        (Self Verification Phase)

Background

     The approach that has been taken by the Organic Chemicals Branch
for the verification phase of the BAT review 1s unique among the
Industrial categories covered by the Effluent Guidelines Division.
To provide the most extensive data set possible on the chemical
Industry, a program has been devised to allow representative plants
to self-sample and analyze their own process streams. By collecting
samples for 30-60 consecutive days, the project officers will develop
a data base representing the variations 1n the waste streams over
a significant period of operating time. The success of this type
of program will depend, 1n a large part however, on the control
that the EPA maintains over the various Industrial laboratories.
Since 1t has become Increasingly obvious that the caliber of quality
assurance programs 1n commercial and Industrial labs ranges widely
1n terms of acceptability, 1t 1s desirable to Initiate a well-controlled
quality assurance program prior to the Initial samplings. This will
Insure that the data received at the conclusion of the program 1s
of known quality and should prevent any need for resampling as a
result of questionable analytical design.

Summary

     To meet this need for a uniform quality assurance procedure, a
program has been developed to provide as much data on the precision
and accuracy of the analyses as 1s possible in the prescribed time
limits, without overburdening the Industrial laboratory. In doing so,
we have attempted to apply the EMSL guidelines of March, 1978.*
Although these were written for a long-term routine analysis program
and required modification for moderate length programs such as
this, they do Incorporate most facets of a basic QA procedure, and
as such have served as a guide for our use.

     The program has assumed that a viable methodology has already
been developed and has been designed to Include a two to three week
period for the Initial work-up of the procedures. During this period
of time, the Instrument capabilities will  be Investigated, validation
of the selected methods will be completed, and Initial precision and
accuracy data will be obtained. This preparation period 1s essential
to Insure that plant personnel are familiar with all procedures
and that data of known, and acceptable, quality can be compiled.

     External standards are best utilized 1n this work to avoid
any possible Interference with the spectrum of the compounds of
Interest. External standards are defined as a known amount of a pure
compound of Interest that 1s measured externally from Identically
the same compound in a sample or sample extract. The con-
centration of the standard should be within a factor of two of the
sample concentration. The recommended method of use 1s to quantltate
          Quality Assurance Program  for  the Analyses of Chemical
          Constituents  in  Environmental  Samples   (March, 1978)


                            JT-  1

-------
•ach run with a standard run Immediately following 1t. In the
Interest of time and sample through-put however, 1t Is acceptable
to use a calibration curve for quantltatlon 1f 1t 1s checked and
updated between every three analyses. If slgnTf leant deviations
occur between standard runs, 1t must be Immediately corrected to
reflect the condition changes.

Time Requirements

     The time required for such a program will be significant, but
adherence to the outlined procedures should allow problems to be
recognized when they first occur and thus avoid any repetition of
sampling or analysis at a later date. The actual percentages of time
for quality assurance are difficult to predict for any one plant without
knowledge of the number of sampling points, the nature of the sampling
points, and the number of data points gained from each method utilized.
The data gained from the Initial two weeks may allow several sampling
points to be grouped for obtaining QA data. Thus the number of spiked
samples and duplicates may well be reduced. The following data on
time needed 1s determined as 1f each point and each method were
unique.

Time Required for N data points (assuming 1 sample per sample point
per day for 30 days and 1 method of analysis)

     N * number of data points gained from each analysis (e.g. N «
     25 1f 25 priority pollutants are characterized by the same
     method.)

     30 (N) » number of data points  (defined above) over 30 days.

Initial Analyses (Two week preparation period)

     (1) 5 (X) analyses for determining linearity of GC where X «
         number of detector/column combinations to be used.

     (2) Distilled Water Method Validation
                   4 extractions
                   8 (X) analyses
                   3 (X) standard analyses for quantltatlon
     (3) Sample Matrix Method  Validation
                   2 extractions for Initial quantltatlon
                   4 (X) analyses for Initial quantltatlon
                   4 extractions for spiked analyses
                   8 (x) analyses for spiked analyses
                   3 (X) standard analyses for quantltatlon
                         for  spiked  analyses

Routine Analyses   (30 day  survey)

     (1) Data  Points

-------
          30 extractions
          30 (x) analyses

     (2) Instrument checks (each shift)
          Composite analytical standards of pollutants 1n question-
               set of standards to check Instrument
               and  update standard curve

     (3) Spiking/Recovery Studies
          10 extractions (1 every 3 days)
          10 (x) analyses

     (4) Duplicate Analyses
               20 extractions (1 every 3 days)
               40 (x) analyses     (Includes duplicates for both
                                   samples and spiking studies)

               for every 30 (x) sample points, 60 extractions
               will be required.

               for every 30 (x) sample points, 120 (x) analyses
               will be required, plus additional runs to keep calibration
               curve update or to quantltate ( 40-120 additional
               analyses).

Recommendations

     The following program of quality assurance should be released
to the participating Industries with the recommendation that 1t be
followed as closely as possible to Insure complete and comparable
data from all laboratories. Alternative plans may be drafted and
utilized by the Industrial firms 1f they are considered to be of
equal merit by the EPA, but these must be well-documented and be
approved prior to use. Industries may be warned that 1f their program
1s judged to.be Inadequate, or scientifically unsound, the data will
not be given full consideration 1n the evaluation of the verification
phase results.

Instrument Requirements and Calibration

     The following equipment Is recommended:
               1. One or more gas chromatographs equipped with
          two different detector types, 1f possible, under a given
          method.
               2. Two distinct column types, preferably one polar
          and one non-polar to allow Independent qualitative and
          quantitative determinations. It 1s necessary, for truly
          reliable characterization to have two Independent con-
          firmations of the data.
               3. An automated Injection system. The accuracy of
          external standards depends upon the consistency of the

-------
          Injection technique coupled with operator control. The
          use of an automated system will standardize that source
          of error and will allow fewer duplicate analyses during
          the routine analysis program.
               4. A data handling system. The volume of data that
          will be generated 1n the relatively short period of time
          will necessitate the use of a data handling program.
          at the least, a peak Integrator.

     Before any GC runs are made for quant1 tat1on purposes, the linearity
of the detectors must be determined. This 1s done by analyzing four or
more solutions of the priority pollutants of Interest. The concentrations
of the pollutants must vary, with two bracketing the range of expected
concentration and the remainder spaced at equal Intervals along the curve.
If the range 1s very narrow, four standards will suffice. Otherwise, five
or more must be used to sufficiently determine linearity. Once the
working range has been established, samples must be diluted 1f necessary,
to remain within the proper limits. These limits must be determined 1n
all the detectors, and must be rechecked 1f any major changes are made
(e.g. replacement of detectors).
     Records should be kept on all Instruments during the period of
preparation and the 30 day sampling period. These records should Include
descriptions of all prepared columns and of all maintenance, routine or
emergency. In addition, the preparation of all standard solutions should
be documented.

Construction of Calibration Curve

     Prior to analysis of samples, a standard curve that covers the
entire working range of the method must be constructed with at least
five standards, Including one near the upper limit of the concen-
tration range. The other standards should be equally spaced throughout
the operating concentration range. The data gained from the determina-
tion of linearity, described above, may be used for this purpose.

     Each day, 1f operation 1s continuous, or prior to analyzing each
group of samples, 1f operation 1s non-continuous, analyze a minimum
of two standards to establish the validity of the original standard
curve. These standards should represent the range of the standard
curve. If these standards fall outside the established limits, a new
standard curve must be constructed. These limits should be established
by the analyst as a part of his ongoing quality control program.

Method Validation; I. Distilled Water Matrix

     To Insure that the prescribed methods are viable for given water
samples, each must be tested through a series of distilled water spikes.
This will also Indicate the ability of the laboratory to follow the
procedures.

     A large sample of distilled water of known purity 1s spiked, and
thoroughly mixed, with those priority pollutants of Interest. These
spikes should be within the ranges observed in the screening phase, or,

-------
If such data 1s not available, it ten times the limit of detection.
Four extractions (and/or purges or appropriate method of preparation)
should be performed according to the specified method, and each should
be analyzed 1n duplicate under each set of conditions (different -
columns, different detectors, etc,}.

     The parameters of Interest must be quantltated using updated
calibration curves. For each parameter, use the resulting observed
values (01, 02, 03, and 04) to calculate the standard deviation
(s) of the replicates as follows:
                    )S
                  \J

                              v>«,  *\**drr o^ f* r*' • f r^f
     Using the same set of data, calculate the mean percent recovery
(P) of the method as follows:
where T « theoretical concentration that should have resulted from
the spike.
     The precision and accuracy criteria should be recorded and sub-
mitted to the Environmental Protection Agency prior to the Initiation
of sample analysis. These Initial criteria are supplemented by, and
periodically revised, updated, and Improved using the data acquired
during the laboratory's routine analytical quality assurance program.

Method Validation: II. Sample Matrix

      For this work, a large sample from each waste stream should be
composited from several smaller samples so that five extractions (or
purges) can be completed. One portion should be prepared and analyzed
with no added fortification so that the amount of each pollutant
already present can be determined. The remaining portions should
then be spiked at levels approximating that of the actual sample,
and four replicate extractions should be made. Each extract should
be analyzed 1n duplicate under each of Instrumental conditions. The
standard deviations of the observed values should be recorded and
preliminary quality control charts should be made as described below.
This data should be reviewed by EPA personnel before the actual
sampling begins.

Construction of Quality Control Charts

     For each pollutant, a chart 1s made showing the standard deviation
of the observed values. The charts allow a quick review of the data
to Insure that the system 1s not out of control.

-------
When the results exceed the boundaries (1n this case 3 standard
deviation units) the system 1s considered to be out of control.
Further work should not be resumed until  the problem(s) 1s Isolated
and resolved.

Procedure for 30 Day Analysis Program

     Once the Initial data Indicates that the laboratory has adequate
control of the system, the routine sampling and analysis may proceed.
Quality control data will be charted as previously described and
any sign of loss of control must be Investigated. The analysis program
must Include the following elements:

     (1) Method blanks
              A method blank must be determined for each set of samples
     analyzed and whenever a new source (new container) of reagent
     or solvent 1s Introduced Into the analytical scheme. (Note:
     the Individual solvents and reagents should be checked for
     purity prior to use 1n determining the method blank or 1n
     the analysis of samples.)

              To determine the method blank, take a quantity of reagents
     equivalent to that used 1n the analysis of the sample and carry
     them through the entire analytical procedure Including all
     glassware and other materials that may come Into contact with
     the sample. Determine a method blank for each class of compounds
     to be determined, I.e., pesticides, base-neutrals and acids,
     metals, phenol1cs, cyanides, etc.

              Reagents having background levels that Interfere with
     the compounds to be determined must be purified and shown to
     be acceptable or replaced with some that are acceptable prior
     to proceeding with the analyses. Problems encountered and
     corrective actions taken should be reported for Information and
     possible resolution of problems encountered by other analysts.

     (2) Field Blanks
              A field blank must be analyzed with each set of samples
     from a given source. This 1s particularly Important whenever
     automatic samplers are used for collection of samples. The blanks
     must be analyzed 1n the same manner as the sample. Field blanks
     for purgeables are sent from the laboratory to the sampling site
     and returned as a check on possible contamination of the sample
     by permeation of volatHes through the septum seal.

     When Interferences occur, the analytical results must be dis-
     carded unless sufficient data from these blanks 1s available to
     permit correction of the results.

-------
(3) Spiked Samples

     The sample from each sampling point (or group of sampling
points) 1s used In spiking and recovery studies every third '
day.Determine the recovery of the method for the analysis of
environmental samples by adding a spike (T;, true value)
sufficient to approximately double the background concentration
level (X-J of the sample selected. If the original concentration
1s higher than the midpoint of the standard curve (range  of the
method), then the concentration of the spike should be
approximately one-half the original concentration. If the con-.
centratlon of the original sample was not detectable, the
concentration of the spike should be five to fifteen times the
lower limit of detections The volume of standard added 1n
aqueous solution should not dilute the sample by more than
ten percent. The volume of standard added In an organic solvent
solution should be kept small (100 ul/1 or less), so that the
solubility of the standard 1n the water will not be affected.

     Analyze the sample, calculate the observed value (Oj
and then calculate the recovery for the spike as follows:
where PJ. 1s the percent recovery. If the sample was diluted due
to the addition of the spike, adjust X-t accordingly.

     After the Initial analyses (See above) the first five
recovery data points should be combined with those determined
on the Initial method validation work and the mean percent
recovery (P) and standard deviation (3f) of the recovery should
be calculated:       .-
                      t»\
where n « the number of percent recovery values available.

Subsequent results should be compared to this data. If the
percent recovery of the spike 1s not within the Interval of
P +  3S? , the system 1s out of control, and the source of this
systematic error should be Identified and resolved before
continuing with routine analysis. Quality control charts, as
previously described should be maintained.

     (4) Replicate Analyses

     Every third sample should be extracted 1n duplicate and
each extract should be analyzed 1n duplicate. After two weeks,
                          IT-7

-------
use the obtained results to calculate  the  range  (R;) of these
results as follows:
where R;1s the difference between  the results of the pair
(x^and xlt) from sample 1 « 1  through  n. The concentration
of each sample 1s represented by the mean:
                                 '
                            a.
      -                      .
where Xi 1s the average of the results of the replicate pair.
A preliminary estimate of the critical difference  (R^) between
replicate analyses for any specific  concentration  level (C) can
be calculated as:
From these data develop a table of such  Revalues for various
C values that span the concentration  range of  Interest.

         These preliminary drtlcal difference values may be used
to Judge the acceptability of the  succeeding replicate results.
To do this, calculate the mean(x)  and difference  (R) between
the replicate results. Referring to the  table  of critical range
values developed above, find the nearest to 3  and use Its R^
to evaluate the acceptability of R. If the R 1s greater than
R*, the system precision 1s out of control and the source of
this unusual variability should be Identified  and resolved before
continuing with routine analyses.

(5) Check Samples

         Check samples should be prepared from standards of known
purity (e.g. compounds received from  National  Bureau of
Standards, or other recognized primary standards). The check
standards should be approximately  equal  to the concentration
found 1n routine samples. Alternately, one standard above and
one standard below the midpoint of the range of the method should
be used. Analyze the standard and  calculate Oc (the observed
value). The percent recovery (P;)  1s  then calculated as follows:
                  ,*  \ooCoi)
                  V     «—••-•—•
where T^ * the true value.

         After determining the Pj, for approximately 15  check standards,
calculate the mean (P) and standard deviation (S^)  of the  percentages
as  follows:
                              - 8

-------
             ?


     and:



             V  J^iU.**-  (£,p)i/lx

     where n « the number of results available.

             If the percent recovery for succeeding check standards  1s
     not within the Interval  of P *_2 Sp, the system should be checked
     for problems. If problems exist, they must  be resolved before
     continuing with routine analysis.

Evaluation of Quality Assurance Data

     After the Initial two weeks of sampling, the precision and accuracy
data should be compiled to represent the Initial five samples.
Standards of performance should be defined based on these results,
and the remaining analyses should fit the defined criteria. Procedures
and equations for obtaining this Information have been given under
the appropriate section. Problems must be resolved before the analysis
1s continued.

     At the end of the program all quality assurance data should be
tabulated and returned to EPA. All data sheets,  Including charts
and raw data should be Included 1n the final report.
                                 jr-9

-------
                    Chapter III




                   PROPOSED PRIORITY POLLUTANT



                         TEST PROCEDURES
Parameter (ug/1)
Method No.
STORET No,
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
acenaphthene
acrolein
aery Ion itrile
benzene
benzidine
carbon tetrachloride
chlorobenzene
1 ,2,4-trichlorobenzene
hexachlorobenzene
1,2-dichloroethane
1 ,1 ,1-trichloroethane
hexachloroethane
1,1-dichloroethane
1 , 1 ,2-trichloroethane
1 , 1 ,2,2-tetrachloroethane
chloroethane
b is (chloromethyl) ether
bis(2-chloroethyl)ether
2-chloroethyl vinyl ether
610,
603,
603,
602,
605,
601,
601,
612,
612,
601,
601,
612,
601,
601,
601,
601,
625
624
624
624
625*
624*
602, 624*
625*
625
624
624
625
624
624*
624*
624
No method
611,
601,
625
624
34205
34210
32415
34030
39120
32102
34301
34551
39700
34531
34506
34396
34496
34511
34516
34311
34268
34273
34576

-------
20.   2-chloronaphthalene
21.   2,4,6-trichlorophenol
22.   4-chloro-3-methyl phenol
23.   chloroform
24.   2-chlorophenol
25.   1,2-dichlorobenzene
26.   1,3-dichlorobenzene
27.   1,4-dichlorobenzene
28.   3,3'-dichlorobenzidine
29.   1,1-dichloroethene
30.   trans-l,2-dichloroethene
31.   2,4-dichlorophenol
32.   1,2-dichloropropane
33.   trans-l,3-dichloropropene
34.   cis-l,3-dichloropropene
35.   2,4-dimethylphenol
36.   2,4-dinitrotoluene
37.   2,6-dinitrotoluene
38.   1,2-diphenylhydrazine
39.   ethylbenzene
40.   fluoranthene
41.   4-chlorophenyl phenyl ether
42.   4-bromophenyl phenyl ether
43.   bis(2-chloroisopropyl)ether
612, 625                34581
604, 625                34621
604, 625                34452
601, 624*               32106
604, 625                34586
601, 602, 612, 625      34536
601, 602, 612, 625      34566
601, 602, 612, 625      34571
605, 625                34631
601, 624                34501
601, 624                34546
604, 625                34601
601, 624                34541
601, 624                34561
601, 624                34561
604, 625                34606
609, 625                34611
609, 625                34626
625                     34346
602, 624                34371
610, 625                34376
611, 625                34641
611, 625                34636
611, 625                34283
                               11

-------
44.   bis(2-chloroethoxy)methane
45.   methylene chloride
46.   chloromethane
47.   bromomethane
48.   bromoform
49.   bromodichloromethane
50.   fluorotrichloromethane
51.   dichlorodifluoromethane
52.   chlorodibromomethane
53.   hexachlorobutadiene
54.   hexachlorocyclopentadiene
55.   isophorone
56.   naphthalene
57.   nitrobenzene
58.   2-nitrophenol
59.   4-nitrophenol
60.   2,4-dinitrophenol
61.   4,6-dinitro-2-methylphenol
62.   N-nitrosodimethylamine
63.   N-nitrosodiphenylamina
64.   N-nitrosodipropylamine
65.   pentachlorophenol
66.   phenol
67.   bis(2-ethylhexyl)phthalate
611, 625
601, 624*
601, 624
601, 624
601, 624
601, 524
601, 624
601
601, 624
612, 625
612, 625
609, 625
610, 625
609, 625
604, 625
604, 625
604, 625
604, 625
607, 625
607, 625
607, 625
604, 625*
604, 625
606, 625
34278
34423
34418
34413
32104
32101
34488
34668
32105
34391
34386
34408
34696
34447
34591
34646
34616
34657
34438
34433
34428
39094
34694
39100

-------
68.   benzyl butyl phthalate
69.   di-n-butyl phthalate
70.   di-n-octyl phthalate
71.   diethyl phthalate
72.   dimethyl phthalate
73.   benzo(a)anthracene
74.   benzo(a)pyrene
75.   benzo (b) fluoranthene
76.   benzo(k)fluoranthene
77.   chrysene
78.   acenaphthylene
79.   anthracene
80.   benzo(ghi)perylene
81.   fluorene
82.   phenanthrene
83.   dibenzo(ah)anthracene
84.   indeno(l,2,3-cd)pyrene
85.   pyrene
86.   tetrachloroethene
87.   toluene
88.   trichloroethene
89.   vinyl chloride
90.   2,3,7,8-tetrachlorodibenzo
      -p-dioxin
606, 625
606, 625
606, 625
606, 625
606, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
610, 625
601, 624
602, 624
601
601

613
34292
34110
34596
34336
34341
34526
34247
34230
34242
34320
34200
34220
34521
34381
34461
34556
34403
34469
34475
34010
39180
39175

34675

-------
91.   aldrin
92.   dieldrin
93.   chlordane
94.   4,4'-DDT
95.   4,4'-OOE
96.   4,4'-ODD
97.   endosulfan I
98.   endosulfan II
99.   endosulfan sulfate
100.  endrin
101.  endrin aldehyde
102.  heptachlor
103.  heptachlor epoxide
104.  a-BHC
105.  b-BHC
106.  d-BHC
107.  g-BHC
108.  toxaphene
109.  PCB-1016
110.  PCB-1221
111.  PCB-1232
112.  PCB-1242
113.  PCB-1248
114.  PCB-1254
115.  PCB-1260
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625
608, 625*
608, 625
608, 625*
608, 625*
608, 625*
608, 625
608, 625
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
608, 625*
39330
39380
39350
39300
39320
39310
34361
34356
34351
39390
34366
39410
39420
39337
39338
39259
39340
39400
34671
39488
39492
39496
39500
39504
39508
*Approved EPA procedures are already available for these parameters
 (Federal Register, December 1, 1976).

-------
                      PURGEABLE HALOCARBONS
                            METHOD 601
1.   Scope and Application
    1.1   This method covers the determination of 29 purgeable
         halocarbons.   The following parameters may be deter-
         mined by this method:
         Parameter                   STORET No.
         Bromoform                   32104
         Bromodichloromethane        32101
         Bromomethane                 34413
         Carbon tetrachloride        32102
         Chlorobenzene               34301
         Chloroethane                 34311
         2-Chloroethylvinyl ether    34576
         Chloroform                  32106
         Chloromethane               34418
         Dibromochloromethane        34105
         1,2-Dichlorobenzene         34536
         1,3-Dichlorobenzene         34566
         1,4-Dichlorobenzene         34571
         Dichlorodifluoromethane     34668
         1,1-Dichloroethane          34496
         1,2-Dichloroethane          34531
         1,1-Dichloroethene          34501
         trans-l,2-0ichloroethene    34546

                        771     I

-------
     Parameter                   Storet No.
     1,2-Dichloropropane         34541
     cis-l,3-0ichloropropene     34561
     trans-l,3-Dich1oropropene   34561
     Methylene chloride          34423
     1,1,2,2-Tetrachloroethane   34516
     Tetrachloroethene           34475
     1,1,1-Trichloroethane       34506
     1,1,/2-Trichloroethane      34511
     Trichloroethene             39180
     Trichlorofluoromethane      34488
     Vinyl chloride              39175
1.2  This method is applicable to the determination of these
     compounds in municipal and industrial  discharges.  It
     is designed to be used to meet the monitoring require-
     ments of the National Pollutant Discharge Elimination
     System (NPDES).  As such, it presupposes a high expec-
     tation of finding the specific compounds of interest.
     If the user is attempting to screen samples for any or
     all of the compounds above,  he must develop independent
     protocols for the verification of  identity.
1.3  The sensitivity of this method is  usually dependent
     upon the level of interferences rather than instrumen-
     tal limitations.  The limits of detection listed in
     Table 1 represent sensitivities that can be achieved in
     wastewaters under optimum operating conditions.

-------
    1.4  This method is recommended for use only by experienced
         residue analysts or under the close supervision of such
         qualified persons.
2.  Summary of Method
    2.1  An inert gas is bubbled through a 5 ml water sample
         contained in a specially-designed purging chamber.  The
         halocarbons are efficiently transferred from the
         aqueous phase to the vapor phase.  The vapor is swept
         through a short sorbent tube where the halocarbons are
         trapped.  After the purge is completed, the trap is
         heated and backflushed with gas to desorb the halo-
         carbons into a gas chromatographic system.  A temper-
         ature program is used in the GC system to separate the
         halocarbons before detection with a halide-specific
         detector.
    2.2  If interferences are encountered, the method provides
         an optional gas chromatographic column that may be
         helpful in resolving the compounds of interest from the
         interferences.
3.  Interferences
    3.1  Impurities in the purge gas and organic compounds
         out-gasing from the plumbing ahead of the trap account
         for the majority of contamination problems.  The analy-
         tical system must be demonstrated to be free from
         contamination under the conditions of the analysis by
                                 3

-------
     running method blanks.  Method blanks are run by charg-
     ing the purging device with organic-free water and
     analyzing it in a normal manner.  The use of non-TFE
     plastic tubing, non-TFE thread sealants, or flow
     controllers with rubber components in the purging
     device should be avoided.
3.2  Samples can be contaminated by diffusion of "volatile
     organics (particularly freons and methylene chloride)
     through the septum seal into the sample during shipment
     and storage.  A sample blank prepared from organic-free
     water and carried through the sampling and handling
     protocol can serve as a check on such contamination.
3.3  Cross contamination can occur whenever high level and
     low level samples are sequentially analyzed.  To reduce
     the likelihood of this, the purging device and sample
     syringe should be rinsed out twice between samples with
     organic-free water.  Whenever an unusually concentrated
     sample is encountered,  it should be followed by an
     analysis of organic-free water to check for cross
     contamination.  For samples containing large amounts of
     water-soluble materials, suspended solids, high boiling
     compounds or  high organohalide levels, it may be
     necessary to wash out the purging device with a soap
     solution, rinse with distilled water, and then dry in a
     105°C oven between analyses.

-------
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete sampling.
         4.1.1  Vial, with cap - 40 ml capacity screw cap
                (Pierce f13075 or equivalent).  Detergent wash
                and dry at 105°C before use.
         4.1.2  Septum - Teflon - faced silicone (Pierce #12722
                or equivalent).  Detergent wash, rinse with tap
                and distilled water, and dry at 105°C for one
                hour before use.
    4.2  Purge and trap device - The purge and trap equipment
         consists of three separate pieces of apparatus:  the
         purging device, trap, and desorber.  Several complete
         devices are now available commercially.  The device
         must meet the following specifications:  the unit must
         be completely compatible with the gas chromatographic
         system; the purging chamber must be designed for a 5 ml
         volume and be modeled after Figure 1; the dimensions
         for the sorbant portion of the trap must meet or exceed
         those in Figure 2.  Figures 3 and 4 illustrate the
         complete system in the purge and the desorb mode.
    4.3  Gas chromatograph - Analytical system complete with
         programmable gas chromatograph suitable fcr on-column
         injection and all  required accessories Including
         halide-specific detector,  column supplier,  reorder,
         and gases.  A data system for measuring -     areas is
         recommended.

-------
    4.4  Syringes - 5-rol  glass hypodermic with luerlok tip (2
         each).
    4.5  Micro  syringes - 10,  25, 100 ul.
    4.6  2-way  syringe valve with Luer ends (3 each).
    4.7  Syringe - 5-ml gas-tight with shut-off valve.
    4.8  Bottle - 15-ml screw-cap, with Teflon cap liner.
5.  Reagents
    5.1  Sodium thiosulfate -  (ACS) Granular.
    5.2  Trap Materials
         5.2.1   Porous polymer packing 60/80 mesh chroma-
                tographic grade Tenax SC (2,6-diphenylene oxide).
         5.2.2  Three percent  OV-1 on Chromosorfa-W 60/80 mesh.
         5.2.3  Silica gel-(35/60 mesh) - Oavison, grade-15 or
                equivalent.
         5.2.4  Coconut- charcoal 6/10 mesh Bamafay Chaney,
                CA-580-26 lot # M-2649 or equivalent.
    5.3  Activated carbon - Filtrasorb-200 '(Calgon Corp.) or
         equivalent.
    5.4  Organic-free water
         5.4.1  Organic-free water is defined as water free of
                interference when employed in the purge  and trap
                procedure described herein.  It is generated by
                passing tap water through a carbon filter bed
                containing  about  1 Ib. of

-------
            activated carbon.
     5.4.2  A water purification system (Millipore Super-Q
            or equivalent) may be used to generate
            organic-free deionized water.
     5.4.3  Organic-free water may also be prepared by
            boiling water for 15 minutes.  Subsequently,
            while maintaining the temperature at 90°C,
            bubble a contaminant-free inert gas through the
            water for one hour.  While still hot, transfer
            the water to a narrow mouth screw cap bottle and
            seal with a Teflon lined septum and cap.
5.5  Stock standards - Prepare stock standard solutions in
     methyl alcohol using assayed liquids or gas cylinders
     as appropriate.  Because of the toxicity of some of the
     organohalides, primary dilutions of these materials
     should be prepared in a hood.  A NIOSH/MESA approved
     toxic gas respirator should be used when the analyst
     handles high concentrations of such materials.
     5.5.1  Place about 9.8 ml of methyl alcohol into a 10
            ml ground glass stoppered volumetric flask.
            Allow the flask to stand, unstop- pered, for
            about 10 minutes or until all alcohol wetted
            surfaces have dried.   Weigh the flask to the
            nearest 0.1 mg.
                    nr-

-------
5.5.2  Add the assayed reference material:
     5.5.2.1 Liquids - Using a 100 ul syringe, imme-
             diately add 2 drops of assayed reference
             material to the flask, then reweigh.  Be
             sure that the 2 drops fall directly into
             the alcohol without contacting the neck of
             the flask.
     5.5.2.2 Gases - To prepare standards for any of
             the six halocarbons that boil below 30°C
             (bromomethane, chloroethane,
             ch1oromethane, di ch1orodif1uoromethane,
             trichlorofluoromethane, vinyl chloride),
             fill a 5 ml valved gas-tight syringe with
             the reference standard to the 5.0-ml
             mark.  Lower the needle to 5 rran above the
             methyl alcohol menicus.  Slowly inject the
             reference standard above the surface of
             the liquid  (the heavy gas will rapidly
             dissolve into the methyl alcohol).
5.5.3  Reweigh, dilute to volume, stopper, then mix by
       inverting the flask several times.  Transfer the
       standard solution to a 15 ml screw-cap bottle
       with a Teflon cap liner.
5.5.4  Calculate the concentration in micrograms per
       microliter from the net gain in weight.
                  UT-

-------
         5.5.5  Store stock standards at 4°c.  Prepare fresh
                standards weekly for the six gases and
                2-chloroethylvinyl ether.  All other stan- dards
                must be replaced with fresh standard each month.
6.  Calibration
    6.1  Using stock standards, prepare secondary dilution stan-
         dards in methyl alcohol that contain the compounds of
         interest, either singly or mixed together.  The
         standards should be prepared at concentrations such
         that the aqueous standards prepared in 6.2 will com-
         pletely bracket the working range of the analytical
         system.
    6.2  Using secondary dilution standards, prepare calibration
         standards by carefully adding 20.0 ul of standard in
         methyl alcohol to 100, 500, or 1000 ml of organic-free
         water.  A 25 ul syringe (Hamilton 702N or equivalent)
         should be used for this operation.  These aqueous
         standards must be prepared fresh daily.
    6.3  Assemble the necessary gas chromatographic apparatus
         and establish operating parameters equivalent to those
         indicated in Table 1.  By injecting secondary dilution
         standards, establish the sensitivity limit and the
         linear range of the analytical system for each compound.
    6.4  Assemble the necessary purge and trap device.  The trap
         must meet the minimum specifications as shown in

-------
         Figure 2 to achieve satisfactory results.  Condition
         the trap overnight at 180°C by backflushing with an
         inert gas flow of at least 20 ml/nrin.   Prior to use,
         daily condition traps 10 minutes while backflushing at
         180°C.  Analyze aqueous calibration standards (6.2)
         according to the purge and trap procedure in Section
         8.   Compare the responses to those obtained by injec-
         tion of standards (6.3); to determine  purging effici-
         ency and also calculate analytical precision.  The
         purging efficiencies and analytical precision of the
         analysis of aqueous standards must be  comparable to
         data presented by Bellar and Lichtenberg (1978) before
         reliable sample analysis may begin.
    6.5  By analyzing calibration standards, establish the sen-
         sitivity limit and linear range of the entire analyti-
         cal system for each compound.
7.  Quality Control
    7.1  Before processing any samples, the %alyst should daily
         demonstrate through the analysis of an organic-free
         water method blank that the entire analytical system is
         i nterference-free.
    7.2  Standard quality assurance practices should be used
         with this method.  Field replicates should be collected
         to validate the precision of the sampling technique.
         Laboratory replicates should be analyzed to validate
         the precision of the analysis.  Fortified samples

                            or   10

-------
         should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification
         of a peak on the gas chromatogram, confirmatory tech-
         niques such as mass spectroscopy should be used.
    7.3  The analyst should maintain constant surveillance of
         both the performance of the analytical system and the
         effectiveness of the method in dealing with each sample
         matrix by spiking each sample, standard and blank with
         surrogate halocarbons.  A combination of bromo-
         chloromethane, 2-bromo-l-chloropropane, and
         1,4-dichlorobutane is recommended to encompass the
         boiling range covered by this method.  From stock
         standard solutions prepared as above, add a volume to
         give 1000 ug of each surrogate to 45 ml of organic-free
         water contained in a 50-ml volumetric flask, mix and
         dilute to volume (20 ng/ul).  Dose 5.0 ul of this
         surrogate spiking solution directly into the 5 ml
         syringe with every sample and reference standard
         analyzed.  Prepare a fresh surrogate spiking solution
         on a weekly basis.
8.  Sample Collection, Preservation, and Handling
    8..1  Grab samples must be collected in glass containers
         having a total volume in excess of 40 ml.  Fill the
         sample bottles in such a manner that no air bubbles
         pass through the sample as the bottle is being filled.
         Seal the bottle so that no air bubbles are entrapped

-------
         in it.   Maintain the hermetic seal on the sample bottle
         until  time of analysis.
    8.2  The samples must be iced or refrigerated from the time
         of collection until extraction.  If the sample contains
         free or  combined chlorine, add sodium thiosulfate
         preservative (10 mg/40 ml will suffice for up to 5 ppm
         Cip) to the empty sample bottles just prior to ship-
         ping to the sampling site, fill with sample just to
         overflowing, seal the bottle, and shake vigorously for
         1 minute.
    8.3  All samples must be analyzed within 14 days of collec-
         tion.
9.  Sample Extraction and Gas Chromatography
    9.1  Adjust the purge gas (nitrogen or helium) flow rate to
         40 ml/min.  Attach the trap inlet to the purging de-
         vice,  and set the device to purge.  Open the syringe
         valve located on the purging device sample introduction
         needle.
    9.2  Remove the plunger from  a 5 ml syringe and attach a
         closed syringe valve.  Open the sample bottle (or stan-
         dard)  and carefully pour the water  into the syringe
         barrel until it overflows.  Replace the syringe plunger
         and compress the sample.  Open the  syringe valve and
         vent any residual  air while adjusting the sample volume
         to 5.0 ml.  Since  this process of taking an aliquot de-
         stroys the validity  :" the sample for future analysis,

-------
     the analyst should fill a second syringe at this time
     to protect against possible loss of data.  Add 5.0 ul
     of the surrogate spiking solution (7.3) through the
     valve bore, then close the valve.
9.3  Attach the syringe-syringe valve assembly to the
     syringe valve on the purging device.  Open the syringe
     valves and inject the sample into the purging chamber.
9.4  Close both valves and purge the sample for 11.0 + .05
     minutes.
9.5  After the 11 minute purge time, attach the trap to the
     chromatograph, and adjust the device to the desorb
     mode.  Introduce the trapped materials to the GC column
     by rapidly heating the trap to 180°C while back-
     flushing the trap with an inert gas between 20 and 60
     ml/min for 4 minutes.  If rapid heating cannot be
     achieved, the gas chromatographic column must be used
     as a secondary trap by cooling it to 30°C (or sub-
     ambient, if problems persist) instead of the initial
     program temperature of 45°C.
9.6  While the trap is being desorbed into the gas chroma-
     tograph, empty the purging chamber using the sample
     introduction syringe.  Wash the chamber with two 5 ml
     flushes of organic-free water.
9.7  After desorbing the sample for approximately four
     minutes recondition the trap by returning the purge and
     trap device to the purge mode.  Wait 15 seconds
                      Jt  •  13

-------
         then  close the syringe valve on the purging device to
         begin gas  flow through the trap.   Maintain the trap
         temperature at 180°C.   After approximately seven
         minutes  turn off the trap heater and open the syringe
         valve to stop the gas flow through the trap.   When cool.
         the  trap is ready for the next sample.
    9.8  Table 1  summarizes some recommended gas chromatographic
         column materials and operating.conditions for the
         instrument.  Included in this table are estimated
         retention  times and sensitivities that should be
         achieved by this method.  An example of the separation
         achieved by column 1 is shown in Figure 5.  Calibrate
         the  system daily by analysis of a minimum of three
         concentration levels of calibration standards.
10.  Calculations
    10.1 Determine the concentration of individual compounds
         directly from calibrations plots of concentration
         (ug/1) vs. peak height or area units.
    10.2 Report results in micrograms per liter.  When duplicate
         and  spiked samples are analyzed, all data obtained
         should be reported.
11.  Accuracy  and Precision
    The U.S.  EPA Environmental Monitoring and Support Laboratory
    in Cincinnati is in the process of conducting an inter-
    laboratory method study to determine the  accuracy and pre-
    cision of this test procedure.

-------
                           BIBLIOGRAPHY
1.  Bellar, T. A., and J. J. Lichtenberg, Journal American Water
    Works Association, Vol. 66, No. 12, Dec. 1974, pp. 739-744.
2.  Bellar, T. A., and J. J. Lichtenberg, "Semi-Automated
    Headspace Analysis of Drinking Waters and Industrial Waters
    for Purgeable Volatile Organic Compounds," Proceeding from
    ASTM Symposium on Measurement of Organic Pollutants in Water
    and Wastewater, June 1978 (In Press).
3.  "Development and Application of Test Procedures for Specific
    Organic Toxic Substances in Wastewaters.  Category 11 -
    Purgeables and Category 12 - Acrolein, Acrylonitrile, and
    Dichlorodifluoromethane."  Report for EPA Contract
    68-03-2635 (In preparation).
                      jzr

-------
                             TABLE 1

         Organohalides Tested Using Purge and Trap Method

                            Retention Time     Detection Limitt
Compound                        (min.)               ug/1
                           Column 1   Column 2
Chloromethane
Bromomethane
Dichlorodifluoromethane
Vinyl chloride
Chloroethane
Methylene chloride
Trichlorofluoromethane
1,1-Dichloroethene
1,1-Dichloroethane
trans-l,2-Dichloroethene
Chloroform
1,2-Dichloroethane
1,1,1 -Tr i ch 1 oroet hane
Carbon tetrachloride
Bromodi chloromethane
1 ,2-Dichloropropane
trans-l,3,-Dichloropropene
Trichloroethene
Di bromoch 1 oromet hane
1,1, 2-Tri chl oroethane
Cis-l,3-dichloropropene
2-Chloroethylvinyl ether
1.50
2.17
2.62
2.67
3.33
5.25
7.18
7.93
9.30
10.1
10.7
11.4
12.6
13.0
13.7
14.9
15.2
15.8
16.5
16.5
16.5
18.0
5.28
7.05
nd
5.28
8.68
10.1
nd
7.72
12.6
9.38
12.1
15.4
13.1
14.4
14.6
16.6
16.6
13.1
16.6
18.1
18.0
nd
0.009
0.03
0.03
0.01
0.01
0.01
0.01
0.006
0.004
0.006
0.006
0.006
0.005
0.007
0.006
0.004
0.006
0.005
0.01
0.006
0.008
0.06

-------
                       TABLE 1 (continued)
         Organohalides Tested Using Purge and Trap Method
                            Retention Time     Detection Limit*
Compound                        (min.)               ug/1
                           Column 1   Column 2
Bromoform
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Chlorobenzene
1 , 3-Dichlorobenzene
1,2-Di Chlorobenzene
1,4-Dichlorobenzene
19.2
21.6
21.7
24.2
34.0
34.9
35.4
19.2
nd
15.0
18.8
22.4
23.5
22.3
0.02
0.006
0.007
0.03
0.04
0.04
0.04
nd = not determined

Column 1 conditions:  Carbopack B 60/80 mesh coated with 1£
    SP-1000 packed in an 8 ft x 0.1 in ID stainless steel or
    glass column with helium carrier gas at 40 ml/min flow
    rate.  Column temperature held at 45°C for 3 min. then
    programmed at 8°C/min. to 220° then held for 15 min.
Column 2 conditions:  Porisil-C 100/120 mesh coated with
    n-octane packed in a 6 ft x 0.1 in ID stainless steel or
    glass column with helium carrier gas at 40 ml/min flow
    rate.  Column temperature held at 50°C for 3 min then
    programmed at 6 C/min to 170  then held for 4 min.
* Detection limit is calculated from the minimum detectable GC
    response being equal  to five times the GC background noise,
    using a Hall  Model  700-A Detector.
                     3D"-  1*7

-------
   OPTIONAL
   FOAM
   TRAP
'/4 IN.
0. D. EXIT
-EXIT % IN.
    0. D.
                  —-14MM 0. D.
                  INLET tt IN.
                       0. D.
    SAMPLE INLET

    2-WAY  SYRINGE VALVE
    •17CM.  20 GAUGE SYRINGE NEEDLE

 -^SMM. 0. D. RUBBER SEPTUM
                              . 0. D.
                           INLET
                           1/4 IN. 0. D.
                    1/16 IN. O.D.
                    ' STAINLESS STEEL
   10MM GLASS FRIT
   MEDIUM POROSITY
                                       13X MOLECULAR
                                       SIEVE PURGE
                                       GAS FILTER
                                         PURGE GAS
                                         FLOW
                                         CONTROL
             Figure 1. Purging device
       izr-M-7

-------
     PACKING PROCEDURE
                  CONSTRUCTION
    GLASS
    WOOL
ACTIVATED., ,„.,.
CHARCOAL 7.7CM[
5MMJ3

   [-•'I
 GRADE 15
          7.7CM
 SILICA GEL
   TEN AX  7.7 CM
3°oOV-1       ,
GLASS WOOL10""»
          5MM
     7-n-XFOOT
   RESISTANCE
 WIRE WRAPPED
     SOLID<
(DOUBLE LAYER)
                         15CM
           7^/FOOT.
         RESISTANCE
        WIRE WRAPPED
              SOLID
       (SINGLE LAYER)
               8CM-
             TRAP INLET
 COMPRESSION
"FITTING NUT
 AND FERRULES

  THERMOCOUPLE/
  CONTROLLER
  SENSOR
                          ELECTRONIC
                          TEMPERATURE
                          CONTROL
                          AND
                          PYROMETER
                        TUBING 25CM
                        0.105 IN. I.D.
                        0.125 IN. O.D.
                        STAINLESS STEEL
      Figure 2. Trap packings and construction to include
               desorb capability
             -or- \i

-------
                                       ®
                                       •c
                                       o

                                       E
                                       09
                                       03

                                       3
                                       a

                                       09
                                       U
                                      •a
                                       a
                                      •o
                                       c
                                       (0
                                       09

                                       E?
                                       3
                                       a
                                       03

                                       £
                                       0)

                                      "o
                                      CO


                                      ri

                                       Q)
IT/-"

-------
         CARRIER GAS
         FLOW CONTROL
   PRESSURE
   REGULATOR
LIQUID INJECTION PORTS
PURGE GAS
FLOW CONTROL
  13X MOLECULAR^
  SIEVE FILTER   ^
                I
              __
    ciT- ~   ~ ' |
      OPTIONAL 4-PORT COLUMN
      SELECTION VALVE
,16-PORT  TRAP INLET
         RESISTANCE WIRE
COLUMN OVEN
 CONFIRMATORY COLUMN
 i DETECTOR
 •ANALYTICAL COLUMN
 i VALVE
              TRAP
             180°C
                                           ON
                                                 HEATER
                                                 CONTROL
                           _y
                                  PURGING
                                  DEVICE
                      Note:
                      ALL LINES BETWEEN
                      TRAP AND GC
                      SHOULD BE HEATED
                      TO 80°C.
Figure 4. Schematic of purge and trap device - desorb mode
                             \1

-------
00 CO
  gs O

  ^E
CC •  l—i
< wo

    "
o s o
o * r».

T «? j
Q. (J — '
O   Q
       3l\!VHi30a01HOVai3i-2 7 'I '
                               •HM

           3i\i\/doadoaoiHOiai-s 7'
       3l\IVH130a01H3iai -L'L'L






      3IM3Hi3a01HOia -2'L-s





      3W3H130a01HOia-L'L
      3IMVHi3IAIOa01HD
                           3IMVX3HOH01HO-L

                       3W3ZN380a01HO
3NVH130a01HOVai31-3' I' L' L

                      3i\ivHi30iAioaaia-3
        3IM3dOadOa01H3ia-C 'L -

-------
                       PURGEABLE AROMATICS
                            METHOD 602

1.   Scope and Application
    1.1  This method covers the determination of various purge-
         able aromatics.  The following parameters may be deter-
         mined by this method:
         Parameter                                  STORET No.
         Benzene                                     34030
         Chlorobenzene                               34301
         1,2-Dichlorobenzene                         34536
         1,3-DiChlorobenzene                         34566
         1,4-Dichlorobenzene                         34571
         Ethylbenzene                                34371
         Toluene                                     34010
    1.2  This method is applicable to the determination of these
         compounds in municipal and industrial discharges.  It
         is  designed to be used to meet the monitoring require-
         ments of the National Pollutant Discharge Elimination
         System (NPDES).  As such, it presupposes a high expec-
         tation of finding the specific compounds of interest.
         If  the user is attempting to screen samples for any or
         all  of the compounds above,  he must develop independent
         protocols for the verification of identity.
    1.3  The  sensitivity of this method is usually dependent
         upon the level of interferences rather than instru-

-------
         mental limitations.  The limits of detection listed in
         Table 1 represent sensitivities that can be achieved in
         wastewaters under optimum operating conditions.
    1.4  This method is recommended for use only by experienced
         residue analysts or under the close supervision of such
         qualified persons.
2.  Summary of Method
    2.1  An inert gas-is bubbled through a 5 ml water sample
         contained in a specially-designed purging chamber.  The
         aromatics are efficiently transferred from the aqueous
         phase to the vapor phase.  The vapor is swept through a
         short sorbent tube where the aromatics are trapped.
         After the purge is completed, the trap is heated and
         backflusned with gas to desorb the aromatic compounds
         into a gas chromatographic system.  A temperature
         program is used in the GC system to separate the
         aromatics before detection with a photoionization
         detector.
3.  Interferences
    3.1  Impurities in the purge gas and organic compounds
         out-gasing from the plumbing ahead of the trap account
         for the majority of contamination problems.  The  ana-
         lytical system must be demonstrated to be free from
         interferences under the conditions of the analysis by
         running method blanks.  Method blanks are run by
         charging the purging device with organic-free water and
                       Tf

-------
         analyzing it in a normal manner.  The use of non-TFE
         plastic tubing, non-TFE thread sealants or flow con-
         trollers with rubber components in the purging device
         should be avoided.
    3.2  Samples can be contaminated by diffusion of volatile
         organics through the septum seal into the sample during
         shipment and storage.  A sample blank prepared from
         organic-free water and carried through the sampling and
         handling protocol can serve as a check on such contam-
         ination.
    3.3  Cross contamination can occur whenever high level and
         low level samples are sequentially analyzed.  To reduce
         the likelihood of this, the purging device and sample
         syringe should be rinsed out twice between samples with
         organic-free water.  Whenever an unusually concentrated
         sample is encountered, it should be followed by an
         analysis of organic-free water to check for cross
         contamination.  For samples containing large amounts of
         water soluble materials, suspended solids, high boiling
         compounds or  high levels of aromatics, it may be
         necessary to wash out the purging device with a soap
         solution, rinse with distilled water, and then dry in a
         105°C oven between analyses.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete sampling.
         4.1.1     Vial, with cap - 40 ml capacity screw cap
                     -BT-  z-o

-------
     (Pierce 113075 or equivalent).  Detergent wash and dry
     at 105°C before use.
     4.1.2  Septum - Teflon - faced silicone (Pierce #12722
            or equivalent).  Detergent wash, rinse with tap
            and distilled water, and dry at 105°C for one
            hour before use.
4.2  Purge and trap device - The purge and trap equipment
     consists of three separate pieces of apparatus:  the
     purging device, trap, and desorber.  Several complete
     devices are available commercially.  The device must
     meet the following specifications:  the unit must be
     completely compatible with the gas chromatographic
     system; the purging chamber must be designed for a 5 ml
     volume and be modeled after Figure 1; the dimensions
     for the sorbant portion of the trap must meet or exceed
     those in Figure 2.  Figures 3 and 4 illustrate the
     complete system in the purge and the desorb mode.
4.3  Gas chromatograph - Analytical system complete with
     programmable gas chromatograph suitable for on-column
     injection and all required accessories including Model
     PI-51-02 photoiom'zation detector (h-nu Systems, Inc.),
     column supplies, recorder, and gases.  A data system
     for measuring peak areas is recommended.
4.4  Syringes - 5-ml glass hypodermic with luerlok tip (2
     each).
4.5  Micro syringes -  10,  25, 100 ul.

-------
    4.6  2-way syringe valve with Luer ends (3 each).
    4.7  Bottle - Ib-ml screw-cap, with Teflon cap liner.
5.  Reagents
    5.1  Sodium thiosulfate - (ACS) Granular.
    5.2  Trap Materials
         5.2.1     Porous polymer packing 60/80 mesh chromato-
                   graphic grade Tenax GC (2,6-diphenylene
                   oxide).
         5.2.2     Three percent OV-1 on Chromosorb-W 60/80 mesh.
    5.3  Activated carbon - Filtrasorb-200 (Calgon Corp.) or
         equivalent.
    5.4  Organic-free water
         5.4.1     Organic-free water is defined as water free
                   of interference when employed in the purge
                   and trap procedure described herein.  It is
                   generated by passing tap water through a
                   carbon filter bed containing about 1 Ib. of
                   activated carbon.
         5.4.2     A water purification system (Millipore
                   Super-Q or equivalent) may be used to
                   generate organic-free deionized water.
         5.4.3     Organic-free water may also be prepared by
                   boiling water for 15 minutes.  Subsequently,
                   while maintaining the temperature at 90°C,
                   bubble a contaminant-free inert gas through
                   the water for one hour.  While still hot,

                    IDT  -   27.

-------
               transfer the water to a narrow mouth screw
               cap bottle and seal with a Teflon lined
               septum and cap.
5.5  Stock standards - Prepare stock standard solutions in
     methyl alcohol using assayed liquids.  Because benzene
     and 1,4-dichlorobenzene are suspected carcinogens,
     primary dilutions of these compound should be prepared
     in a hood.
     5.5.1     Place about 9.8 ml of methyl alcohol into a
               10 ml ground glass stoppered volumetric
               flask.  Allow the flask to stand, unstop-
               pered, for about 10 minutes or until all
               alcohol wetted surfaces have dried.  Weigh
               the flask to the nearest 0.1 mg.
     5.5.2     Using a 100 ul syringe, immediately add 2
               drops of assayed reference material to the
               flask, then reweigh.  Be sure that the 2
               drops fall directly into the alcohol without
               contacting the neck of the flask.
     5.5.3     Dilute to volume, stopper, then mix by in-
               verting the flask several times.  Transfer
               the standard solution to a 15 ml  screw-cap
               bottle with a Teflon cap liner.
     5.5.4     Calculate the concentration  in micrograms per
               microliter from the net gain in weight.
                .TIT-  2-3

-------
         5.5.5     Store stock standards at 4°C.  All stan-
                   dards must be replaced with fresh standard
                   each month.
6.  Calibration
    6.1  Using stock standards, prepare secondary dilution
         standards in methyl alcohol that contain the compounds
         of interest, either singly or mixed together.  The
         standards should be prepared at concentrations such
         that the aqueous standards prepared in 6.2 will com-
         pletely bracket the working range of the analytical
         system.
    6.2  Using secondary dilution standards,,prepare calibration
         standards by carefully adding 20.0 ul of standard in
         methyl alcohol to 100, 500, or 1000 ml of organic-free
         water.  A 25 ul syringe (Hamilton 702N or equivalent)
         should be used for this operation.  These aqueous
         standards must be prepared fresh daily.
    6.3  Assemble the necessary gas chromatographic apparatus
         and establish operating parameters equivalent to those
         indicated in Table 1.  By injecting secondary dilution
         standards, establish the sensitivity limit and the
         linear range of the analytical system for each compound.
    6.4  Assemble the necessary purge and trap device.  The trap
         must meet the minimum specifications shown in Figure 2
         to achieve satisfactory results.  Condition the trap
         overnight at 180°C by backflushing with an inert gas

-------
         flow of at least 20 ml/min.  Prior to use, daily con-
         dition traps 10 minutes while backflushing at 180°C.
         Analyze aqueous calibration standards (6.2) according
         to the purge and trap procedure in Section 8.  Compare
         the responses to those obtained by injection of stan-
         dards (6.3), to determine purging efficiency and also
         to calculate analytical precision.  The purging effici-
         encies and analytical precision of the analysis of
         aqueous standards must be comparable to data presented
         by Bellar and Lichtenberg (1978) before reliable sample
         analysis may begin.
    6.5  By analyzing calibration standards, .establish the
         sensitivity limit and linear range of the entire
         analytical system for each compound.
7.  Quality Control
    7.1  Before processing any samples, the analyst should
         demonstrate daily through the analysis of an
         organic-free water method blank that the entire
         analytical system is interference-free.
    7.2  Standard quality assurance practices should be used
         with this method.  Field replicates should be collected
         to validate the precision of the  sampling technique.
         Laboratory replicates should be analyzed to validate
         the precision of the analysis.  Fortified samples
         should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification
                     77T

-------
         of a peak on the gas chromatogram, confirmatory tech-
         niques such as mass spectroscopy should be used.
    7.3  The analyst should maintain constant surveillance of
         both the performance of the analytical system and the
         effectiveness of the method in dealing with each sample
         matrix by spiking each sample, standard and blank with
         surrogate compounds.
8.  Sample Collection, Preservation, and Handling
    8.1  Collect about 500 ml sample in a clean container.
         Adjust the pH of the sample to about 2 by adding 1:1
         diluted HC1 while stirring vigorously.  If the sample
         contains free or combined chlorine, add 35 mg of sodium
         thiosulfate per part per million of free chlorine per
         liter of sample.  Fill a 40 ml sample bottle in such a
         manner that no air bubbles pass through the sample as
         the bottle is being filled.  Seal the bottle so that no
         air bubbles are entrapped in it.  Maintain the hermetic
         seal on the sample bottle until time of analysis.
    8.2  The samples must be iced or refrigerated from the time
         of collection until extraction.
    8.3  All samples must be analyzed within 7 days of collec-
         tion.
9.  Sample Extraction and Gas Chromatography
    9.1  Adjust the purge gas (nitrogen or helium) flow rate to
         40 ml/min.  Attach the trap inlet to the purging
         device, and set the device to purge.  Open the syringe
                     3ZT-

-------
     valve located on the purging device sample introduction
     needle.
9.2  Remove the plunger from a 5 ml syringe and attach a
     closed syringe valve.  Open the sample bottle (or
     standard) and carefully pour the water into the syringe
     barrel until it overflows.  Replace the syringe plunger
     and compress the sample.  Open the syringe valve and
     vent any residual air while adjusting the sample volume
     to 5.0 ml.  Since this process of taking an aliquot
     destroys the validity of the sample for future analy-
     sis, the analyst should fill a second syringe at this
     time to protect against possible loss of data.  Add the
     surrogate spiking solution  (7.3) through the valve
     bore, then close the valve.
9.3  Attach the syringe-syringe  valve assembly to the
     syringe valve on the purging device.  Open the syringe
     valves and inject the sample into the purging chamber.
9.4  Close both valves and purge the sample for 12.0 +; .05
     minutes.
9.5  After the 12 minute purge time, disconnect the purge
     chamber from the trap.  Dry the trap by maintaining a
     flow rate of 40 cc/min dry  purge gas for 6 min.  Attach
     the trap to the chromatograph, and adjust the device to
     the desorb mode.  Introduce the trapped materials to
     the GC column by rapidly  heating the trap to  180°C
     while backflushing the trap with an inert gas between

                          2-7

-------
     20 and 60 ml/min for 4 minutes.  If rapid heating can-
     not be achieved, the gas chromatographic column must be
     used as a secondary trap by cooling it.to 30°C (or
     subambient, if problems persist) instead of the initial
     program temperature of 50°C.
9.6  While the trap is being desorbed into the gas chroma-
     tograph, empty the purging chamber using the sample
     introduction syringe.  Wash the chamber with two 5 ml
     flushes of organic-free water.
9.7  After desorbing the sample for approximately four
     minutes recondition the trap by returning the purge and
     trap device to the purge mode.  Wait 15 seconds then
     close the syringe valve on the purging device to begin
     gas flow through the trap.  Maintain the trap temper-
     ature at 180°C.  After approximately seven minutes
     turn off the trap heater and open the syringe valve to
     stop the gas flow through the trap.  When cool the trap
     is ready for the next sample.
9.8  Table 1 summarizes the recommended gas chromatographic
     column material and operating conditions for the
     instrument.  Included in this table are estimated
     retention times and sensitivities that should be
     achieved by this method.  An example of the separation
     achieved by this column is shown in Figure 5.  Cali-
     brate the system daily by analysis of a minimum of
     three concentration levels of calibration standards.
                   jr-

-------
10. Calculations
    10.1 Determine the concentration of Individual compounds
         directly from calibrations plots of concentration
         (ug/1) vs. peak height or area units.
    10.2 Report results 1n micrograms per liter.  When duplicate
         and spiked samples are analyzed, all data obtained
         should be reported.
11. Accuracy and Precision
    The U.S. EPA Environmental Monitoring and Support Laboratory
    in Cincinnati is in the process of conducting an interlabor-
    atory method study to determine the accuracy and precision
    of this test procedure.
                      TOT

-------
                           BIBLIOGRAPHY
1.  Bellar, T. A., and 0. J. Lichtenberg, Journal American Water
    Works Association, Vol. 66, No. 12, Dec. 1974, pp. 739-744.
2.  Bellar, T. A., and 0. J. Lichtenberg, "Semi-Automated
    Headspace Analysis of Drinking Waters and Industrial Waters
    for Purgeable Volatile Organic Compounds," Proceeding from
    ASTM Symposium on Measurement of Organic Pollutants in Water
    and Wastewater, June 1978 (In Press).
3.  Bellar, T. A., and J. J. Lichtenberg, "The Determination of
    Purgeable Aromatic Compounds in Drinking Waters and
    Industrial Wastes," (In preparation).
4.  "Development and Application of Test Procedures for Specific
    Organic Toxic Substances in Wastewaters.  Category 11 -
    Purgeables and Category 12 - Acrolein,  Acrylonitrile, and
    Dichlorodifluoromethane."  Report for EPA Contract
    68-03-2635 (In preparation).
                    or-   50

-------
                             TABLE 1
     Chromatography of Aromatics Using Purge and Trap Method
                      Retention Time (min.)     Detection Limit
Compound                                              ug/1
                          Column 1
Benzene
Toluene
Ethyl benzene
Chlorobenzene
1 , 4-Di ch 1 orobenzene
1 , 3-Di ch 1 orobenzene
1,2-Oi Chlorobenzene
3.33
5.75
8.25
9.17
16.8
18.2
25.9
n.d.
n.d.
n.d.
n.d.
n.d.
n.d.
n.d.
nd = not determined

Column 1 conditions: Supelcoport 100/120 mesh coated with 5%
    SP-2100 and 1.75% Bentone-34 packed in a 6 ft.  x 0.085 in ID
    stainless steel column with helium carrier gas at 36 cc/min
    flow rate.  Column temperature held at 50°C for 2 min.
    then programmed at 6 C/min. to 90 C for a final hold.

 Detection limit is calculated from the minimum detectable GC
    response being equal to five times the GC background noise,
    using a h-nu Model PI-51-02 photoionization detector with a
    10.2 ev lamp.
                      Jfl-

-------
   [OPTIONAL
   ! FOAM
   TRAP
EXIT '/4 IN.
    0. D.
                  — 14MM 0. D.
                   INLET V4 IN.
                       .0. D.
'4 IN.
0. D. EXIT
   SAMPLE INLET

   2-WAY SYRINGE VALVE
   17OA. 20 GAUGE SYRINGE NEEDLE

   6MM. 0. D.  RUBBER SEPTUM

       10MM. 0. D.    1/16 IN. O.D.
                  /STAINLESS STEEL
                              IN. 0. D.
                                        13X MOLECULAR
                                        SIEVE PURGE
                                        GAS FILTER
   10MM GLASS FRIT
   MEDIUM POROSITY
                                          PURGE GAS
                                          FLOW
                                          CONTROL
             Figure 1. Purging device
          W

-------
      PACKING PROCEDURE
     GLASS
     WOOL
    TENAX 23CM
i
    3%OV-1lCMj
GLASS WOOL    '
CONSTRUCTION
       COMPRESSION FITTING
       NUT AND FERRULES
        14FT.7-H-/FOOT RESISTANCE
            WRAPPED SOLID
                             n
                             -L
                  THERMOCOUPLE/
                  CONTROLLER
                 • SENSOR
                     ELECTRONIC
                     TEMPERATURE
                     CONTROL
                     AND
                     PYROMETER
                  TUBING 25CM.
                  0.105 IN. I.D.
                  0.125 IN. O.D.
                  STAINLESS STEEL
              TRAP INLET
 Figure 2. Trap packings and construction to include
          desorb capability
                         31 b

-------
 3
 a
 i
 o
 o
TJ

 a
•o

 (0
 a
 10


 «
£
 O

(0


CO


 £
 3
 CD

-------
         CARRIER GAS
         FLOW CONTROL
   PRESSURE
   REGULATOR
PURGE GAS
FLOW CONTROL
LIQUID INJECTION PORTS
  13X MOLECULAR ^
  S.EVE FILTER
                  -COLUMN OVEN
              	CONFIRMATORY COLUMN
        HH,U-r%  !_} TO DETECTOR
                I  ^-ANALYTICAL COLUMN
       OPTIONAL 4-PORT COLUMN
       SELECTION VALVE
  6-PORT TRAP INLET
  VALVE JRES.STANCEW.RE

                       ^ CONTROL
                               'f>
              TRAP
              TRAP
     FLOW     180°C
           PURGING
           DEVICE
                                             Note:
                                             ALL LINES BETWEEN
                                             TRAP AND GC
                                             SHOULD BE HEATED
                                             TO 80°C.
Figure 4. Schematic of purge and trap device - desorb mode

-------
       LU
       INI
            LU

            2
            LU
          O
          CC
          O
          3
          cc


          9
        LU
                   LU
  INI
  2


"3
Z cc
X

LU

INI  Qj
2  INI
LU  2
CO  ii i
O  co
cc  Q
O  cc

=d  O
                                Q  i
                                   t    1    1    t
0   2   4   6   8  10  12  14  16  18  20  22  24  26  23



                   RETENTION TIME-MINUTES


Figure 5. Gas chromatogram of purgeable aromatics

-------
                    ACROLEIN AND ACRYLONITRILE
                            METHOD 603

1.   Scope and Application
    1.1   This method covers the determination of acrolein and
         acrylonitrile.  The following parameters may be deter-
         mined by this method:
                Parameter                   STORET No.
                Acrolein                      34210
                Acrylonitrile                 32415
    1.2   This method is applicable to the determination of these
         compounds in municipal and industrial discharges.  It
         is designed to be used to meet the monitoring require-
         ments of the National  Pollutant Discharge Elimination
         System (NPDES).  As such, it presupposes a high expec-
         tation of finding the  specific compounds of interest.
         If the user is attempting to screen samples for any or
         all of the compounds above,  he must develop independent
         protocols for the verification of identity.
    1.3   The sensitivity of this method is usually dependent
         upon the level  of interferences rather than instru-
         mental limitations.  The limits of detection listed in
         Table 1 represent sensitivities that can be achieved in
         wastewaters under optimum operating conditions.
    1.4   This method is recommended for use only by experienced
                            3z-

-------
         residue analysts or under the close supervision of such
         qualified persons.
2.  Summary of Method
    2.1  An inert gas is bubbled through a 5 ml water sample
         contained in a specially-designed heated purging
         chamber.  Acrolein and acrylonitrile are transferred
         from the vapor phase to the gaseous phase.  The vapor
         is passed through a short sorbent tube where the com-
         pounds are trapped.  After the extraction is completed,
         the trap is heated and backflushed with gas to desorb
         the compounds into a gas chromatographic system.  A
         temperature program is used in the GC system to
         separate the compounds before detection with a flame
         ioni2ation detector.
3.  Interferences
    3.1  Impurities in the purge gas and organic compounds
         out-gasing from the plumbing ahead of the trap account
         for the majority of contamination problems.  The
         analytical system must be demonstrated to be free from
         interferences under the conditions of the analysis by
         running method blanks.  Method blanks are run by charg-
         ing the purging device with organic-free water and
         analyzing it in a normal manner.  The use of non-TFE
         plastic tubing, non-TFE thread sealants, or flow con-
         trollers with rubber components in the purging device
         should be avoided.

-------
3.2  Samples can be contaminated by diffusion of volatile
     organics (particularly methylene chloride) through the
     septum seal into the sample during shipment and
     storage.  A sample blank prepared from organic-free
     water and carried through the sampling and handling
     protocol can serve as a check on such contamination.
3.3  Cross contamination can occur whenever high-level and
     low level samples are sequentially analyzed.  To reduce
     the likelihood of this, the purging device and sample
     syringe should be rinsed out twice between samples with
     organic-free water.  Whenever an unusually concentrated
     sample is encountered, it should be followed by an
     analysis of organic-free water to check for
     cross-contamination.  For samples containing large
     amounts of water soluble materials, suspended solids,
     high boiling compounds or  high organohalide levels it
     may be necessary to wash out the purging device with a
     soap solution, rinse with distilled water, and then dry
     in a 105°C oven between analyses.
3.4  Interferences are sometimes reduced or eliminated by
     first purging the water sample for 5 minutes at room
     temperature in 9.4.  Then the purge device is rapidly
     heated to 85°C and purged as in 9.4.  With such a
     modification, approximately 5 to 10% of the acryloni-
     trile and a trace of the acrolein in the sample will be

-------
         lost.  Therefore, calibration must be established for
         the compounds under the conditions of this modified
         procedure.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete sampling.
         4.1.1     Vial, with cap - 40 ml capacity screw cap
                   (Pierce #13075 or equivalent).  Detergent
                   wash and dry at 105°C before use.
         4.1.2     Septum - Teflon - faced silicone (Pierce
                   #12722 or equivalent).  Detergent wash, rinse
                   with tap and distilled water, and dry at
                   105°C for one hour before use.
    4.2  Purge and trap device - The purge and trap equipment
         consists of three separate pieces of apparatus:  the
         purging device, trap, and desorber.  The purging device
         should be equipped for heating in the same manner as
         the trap (electrically) or with a circulating water
         jacket.  If electrical heating is used the electrical
         parts must be protected so that water will not drip on
         the conductors, causing dangerous electrical shock or
         shorts.  All temperature parameters must be carefully
         controlled.  Several complete devices are available
         commercially although most are not equipped to heat the
         purging chamber.  The device must meet the following
         specifications:  the unit must be completely compatible
         with the gas chromatographic system; the purging
                   ITT"

-------
         chamber must be  designed  for  a  5 ml  volume  and  be
         modeled after  Figure  1; the dimensions  for  the  sorbant
         portion of the trap must  meet or exceed those in figure
         2.   Figures 3  and  4 illustrate  the  complete system in
         the  purge and  the  desorb  mode.
    4.3   Gas  chromatograph  - Analytical  system complete  with
         programmable gas chromatograph  suitable for on-column
         injection, equipped with  matched columns for dual
         column  analysis  and a differential  flame ionization
         detector.  A nitrogen specific  detector (thermionic  or
         Hall) may be used  if  only acrylonitrile is  to be
         detected.  Required  accessories include:  column
         supplies, recorder,  and gases.   A  data  system for
         measuring peak areas  is recommended.
    4.4   Syringes  - 5-ml  glass hypodermic with luerlok tip  (2
         each).
    4.5   Micro  syringes - 10,  25,  100  ul.
    4.6   2-way  syringe  valve with  Luer ends (3 each).
    4.7   Bottle  -  15-ml screw-cap, with Teflon cap liner.
5.  Reagents
    5.1   Preservatives
         5.1.1      Sodium hydroxide -  (ACS) 10 N in distilled
                  water.
         5.1.2      Sulfuric acid - (ACS).  Mix equal volumes of
                   cone.  H-SO, with distilled water.
         5.1.3     Sodium thiosulfate - (ACS) Granular.
    5.2  Trap absorbent - Porous polymer packing, 50/80 mesh
                       2T-

-------
     chromatographic grade Porapak N.
5.3  Activated carbon - Filtrasorb-200 (Calgon Corp.) or
     equivalent.
5.4  Organic-free water
     5.4.1     Organic-free water is defined as water free
               of interference when employed in the purge
               and trap procedure described herein.  It is
               generated by passing tap water through a
               carbon filter bed containing about 1 Ib. of
               activated carbon.
     5.4.2     A water purification system (Millipore
               Super-Q or equivalent) may be used to
               generate organic-free deionized water.
     5.4.3     Organic-free water may also be prepared by
               boiling water for 15 minutes.  Subsequently,
               while maintaining the temperature at 90°C,
               bubble a contaminant-free inert gas through
               the water for one hour.  While still hot,
               transfer the water to a narrow mouth screw
               cap bottle and sealwith a Teflon lined septum
               and cap.
5.5  Stock standards - Prepare  stock standard solutions
     daily in water using assayed standards.  Because of
     toxicity, primary dilutions of these materials should
     be prepared in a hood.  A NIOSH/MESA approved toxic gas
     respirator should be used when the analyst handles high

-------
         concentrations of the materials.
         5.5.1      Place about 9.8 ml  of water (pH 6.5 to 7.5)
                   into a 10 ml ground glass stoppered volu-
                   metric flask.  Allow the flask to stand,
                   unstoppered, for about  10 minutes or until
                   all  water wetted surfaces have dried.  Weigh
                   the  flask to the nearest 0.1  mg.
         5.5.2      Using a 100 ul  syringe, immediately add 2
                   drops of assayed reference material to the
                   flask, then reweigh.  Be sure that the 2
                   drops fall directly into the water without
                   contacting the  neck of the flask.
         5.5.3      Dilute to volume, stopper, then mix by in-
                   verting the flask several times.  Transfer
                   the  standard solution to a 15 ml screw-cap
                   bottle with a Teflon cap liner.
         5.5.4      Calculate the concentration in micrograms per
                   microliter from the net gain in weight.
6.  Calibration
    6.1  Using  stock standards, prepare secondary dilution stan-
         dards  in water.  The standards should be prepared at
         concentrations such that the aqueous standards prepared
         in 6.2 will completely bracket the working range of the
         chromatographic system.
    6.2  Using secondary dilution standards, prepare calibration
         standards by carefully adding ul  of stock standard
                        -£-

-------
         to 100, 500, or 1000 ml of organic-free water.
    6.3  Assemble the necessary gas chromatographic apparatus
         and establish operating parameters equivalent to those
         indicated in Table 1.  By injecting secondary dilution
         standards, establish the sensitivity limit and the
         linear range of the analytical system for each compound.
    6.4  Assemble the necessary purge and trap device.  The trap
         must meet the minimum specifications as shown in Figure
         2 to achieve satisfactory results.  Condition the trap
         overnight at 180°C by backflushing with an inert gas
         flow of at least 20 ml/min.  Prior to use, daily
         condition traps 10 minutes while backflushing at
         180°C.  Analyze aqueous calibration standards (6.2)
         according to the purge and trap procedure in Section
         9.  Compare the responses to those obtained by injec-
         tion of standards (6.3), to determine purging effici-
         ency and also to calculate analytical precision.  The
         purging efficiencies and analytical precision of the
         analysis of aqueous standards should be 85 + 5% for
         acrolein and 98% + 5% for acrylonitrile.
    6.5  By analyzing calibration standards, establish the sen-
         sitivity limit and linear range of the entire analy-
         tical system for each compound.
7.  Quality Control
    7.1  Before processing any samples, the analyst should
         demonstrate daily through the analysis of an

-------
         organic-free water method blank that the entire analy-
         tical  system is interference-free.
    7.2  Standard quality assurance practices should be used
         with this method.  Field replicates should be collected
         to validate the precision of the sampling technique.
         Laboratory replicates should be analyzed to validate
         the precision of the analysis.  Fortified samples
         should be analyzed to validate the" accuracy of the
         analysis
         Where doubt exists over the identification of a peak on
         the gas chromatogram, confirmatory techniques such as
         mass spectroscopy should be used.
    7.3  The analyst should maintain constant surveillance of
         both the performance of the analytical system and the
         effectiveness of the method in dealing with each sample
         matrix by spiking each sample, standard and blank with
         surrogate compounds.
8.  Sample Collection, Preservation, and Handling
    8.1  Collect about 500 ml sample in a clean container.
         Adjust the pH of the sample to 6^5 to 7.5 by adding 1:1
         diluted hUSO^ or NaOH while stirring vigorously.
         If the sample contains residual chlorine, add 35 mg of
         sodium thiosulfate per part per million of free
         chlorine per liter of sample.  Fill a 40 ml sample
         bottle and seal  the bottle so that no air bubbles are
         entrapped in it.  Maintain the hermetic seal on the
                        jr

-------
         sample bottle until time of analysis.
    8.2  The samples must be'iced or refrigerated at 4°C from
         the time of collection until extraction.
    8.3  All samples must be analyzed within 3 days of collec-
         tion.
9.  Sample Extraction and Gas Chromatography
    9.1  Adjust the helium purge gas flow rate to 2(H1 ml/min
         and the temperature of the purge device to 85°C.
         Attach the trap inlet to the purging device, and set
         the device to purge.  Open the syringe valve located on
         the purging device sample introduction needle.
    9.2  Remove the plunger from a 5 ml syringe and attach a
         closed syringe valve.  Open the sample bottle (or
         standard) and carefully pour the water into the syringe
         barrel until it overflows.  Replace the syringe plunger
         and compress the sample.  Open the syringe valve and
         vent any residual air while adjusting the sample volume
         to 5.0 ml.
    9.3  Attach the syringe-syringe valve assembly to the
         syringe valve on the purging device.  Open the syringe
         valves and inject the sample into the purging chamber.
    9.4  Close both valves and purge the sample for 30.0 + 0.1
         minutes.  Monitor and control the temperature of the
         purge device to obtain 85+l°C.
    9.5  After the 30-minute purge time, attach the trap to tN
         chromatograph, and adjust the device to the desorb

-------
     mode.   Introduce the trapped materials to the GC column
     by rapidly heating the trap to 170°C while back-
     flushing the trap with helium at 45 ml/min for 5
     minutes.  The backflushing time and gas flow rate must
     be carefully reproduced from sample to sample.  During
     backflushing the chromatographic column is held at
     100°C.  Record GC retention time from the beginning
     of desorption.
9.6  While the trap is being desorbed into the gas chroma-
     tograph, empty the purging chamber using the sample
     introduction syringe.  Wash the chamber with two 5 ml
     flushes of organic-free water.
9.7  After desorbing the sample for 5 minutes recondition
     the trap by returning the purge and trap device to the
     purge mode and begin the GC program.  Wait 15 seconds
     then close the syringe valve on the purging device to
     begin gas flow through the trap.  Maintain the trap
     temperature at 170°C.  After approximately seven
     minutes turn off the trap heater and open the syringe
     valve to stop the gas flow through the trap.  When cool
     the trap is ready for the next sample.
9.8  Table 1 summarizes some recommended gas chromatographic
     column materials and operating conditions for the
     instrument.  Included in this table are estimated
     retention times and sensitivities that should be
     achieved by this method.  An example of the  separation
                   _ZT-

-------
         achieved by this column is shown in Figure 5.  Cali-
         brate the system daily by analysis of a minimum of
         three concentrations levels of calibration standards.
10. Calculations
    10.1 Determine the concentration of individual compounds
         directly from calibrations plots of concentration
         (ug/1) vs. peak height or area units.
    10.2 Report results in micrograms per liter.  When duplicate
         and spiked samples are analyzed, all data obtained
         should be reported.
11. Accuracy and Precision
    The U.S. EPA Environmental Monitoring and Support Laboratory
in Cincinnati is in the process of conducting an interlaboratory
method study to determine the accuracy and precision of this
test procedure.

                           BIBLIOGRAPHY

1.  Bellar, T. A., and J. J. Lichtenberg, Journal American Water
    Works Association, Vol. 66, No. 12, Dec. 1974, pp. 739-744.

2.  Bellar, T. A., and J. J. Lichtenberg, "Semi-Automated
    Headspace Analysis of Drinking Waters and Industrial Waters
    for Purgeable Volatile Organic Compounds," Proceeding from
    ASTM Symposium on Measurement of Organic Pollutants in Water
    and Wastewater, June 1978 (In Press).

-------
3.  "Development and Application of Test Procedures for Specific
    Organic Toxic Substances in Wastewaters.  Category 11 -
    Purgeables and Category 12 - Acrolein, Acrylonitrile, and
    Dichlorodlfluoromethane."  Report for EPA Contract
    68-03-2635 (In preparation).

4.  Going, John, et al., "Environmental Monitoring Near
    Industrial Sites-Acrylonltrile," EPA Report No.
    560/6-79-003, 1979.

-------
                             TABLE 1
           Gas Chromatography by Heated Purge and Trap
                                               Detection Limit
Compound            Retention Time (min.)      	ug/1
Acrolein                    7.6                       2
Acrylonitrile               8.9                       1
Column conditions:  Chromosorb 101 80/100 mesh packed 1n a
    6' x 1/8" O.D. stainless steel column with helium carrier
    gas at 45 ml/min flow rate.  Column temperature is held at
    100°C for 5 minutes during trap desorption, then
    programmed at 10°C/min to 140°C and held for 5 minutes.

Detection limit is estimated, based upon the use of a flame
    ionization detector.

-------
   OPTIONAL
   FOAM
   TRAP
-EXIT V» IN.
     0. D.

—-14MM 0. D.

 INLET "< IN.
     0. D.
                   -SAMPLE INLET
% IN.  _
0. D. EXIT
     &-^2-WAY SYRINGE VALVE
     ¥— 17CM. 20 GAUGE SYRINGE NEEDLE
•Mk
           ~10MM. 0. D.    1/16 IN. O.D.
        ^^M^HNLET"
       . 0. D. RUBBER SEPTUM
                         V* IN. 0. D.
                                  y STAINLESS STEEL
                                   13X MOLECULAR
                                   SIEVE PURGE
                                   GAS FILTER
   10MM GLASS FRIT
   MEDIUM POROSITY
                                     PURGE GAS
                                     FLOW
                                     CONTROL
            Figure 1. Purging device
           IT-

-------
   PACKING PROCEDURE
  GLASS
  WOOL
 PORAPAK N
      24CM
GLASS WOOL
        5MM
CONSTRUCTION
          COMPRESSION
          .FITTING NUT
          AND FERRULES
           14FT. 7^ /FOOT
                                      RESISTANCE WIRE
                                      WRAPPED SOLID
                                       THERMOCOUPLE/
                                       CONTROLLER
                                       SENSOR
              ELECTRONIC
              TEMPERATURE
              CONTROL
              AND
              PYROMETER
                                       TUBING 25CM
                                       0.105 IN. l.D.
                                       0.125 IN. O.D.
                                       STAINLESS STEEL
           TRAP INLET

Figure 2. Trap packings and construction to include
         desorb capability

-------
                                         4)
                                         o

                                         E
                                         ®
                                         05

                                         3
                                         Q.
                                          i
                                         0)
                                         O
                                         '>
                                         09
                                         •o
                                         a
                                         CO
                                         T3
                                          C
                                          (0

                                          0)
                                          O)
                                          l_
                                          3
                                          a

                                          "o

                                          o
                                          'S
                                          (Q


                                          03
                                          £
                                          U
                                          CO


                                          cri

                                          
-------
         CARRIER GAS
         FLOW CONTROL
   PRESSURE
   REGULATOR
   LIQUID INJECTION PORTS
                     -COLUMN OVEN
PURGE GAS
FLOW CONTROL^!
 13X MOLECULAR
 SIEVE FILTER""

                                          -ANALYTICAL COLUMN
                 	^CONFIRMATORY COLUMN
   ^Qi  pfHr^-^n100^0™
^  iL^Milr—• -  \  ~  	
  "*"      \OPTIONAL4-PORTCOLUMN
          SELECTION VALVE
     6-PORT TRAP INLET
     VALVE j RESISTANCE WIRE  HEATER
           rt^              "CONTROL
 X    *  i
STRAPS.    TRAP ION
 WLOWi    170°C  .,
                                            Note:
                                            ALL LINES BETWEEN
                                 PURGING    TRAP AND GC
                                 DEVICE     SHOULD BE HEATED
                                            TO 95°C.
Figure 4. Schematic of purge and trap device - desorb mode

-------
 :OLUMN: CHROMOSORB 101
 ROGRAM: 80eC-5 MINUTES.
         8°C< MINUTE TO 1505C
DETECTOR: FLAME IONIZATION
                         o
                         c=
                         CJ
or
U
02468
         RETENTION TIME-MINU":

 Figured. Gas chromatogram -
      ein and acrylonitrile

-------
                             PHENOLS
                            METHOD  604

1.   Scope and Application
    1.1   This method  covers  the  determination of various
                                                       •
         phenolic  compounds.  The following parameters may be
         determined by  this  method:
    Parameter                 STORET No.
    4-Chloro-3-methylphenol      34452
    2-Chlorophenol               34586
    2,4-Dichlorophenol           34601
    2,4-Dimethylphenol           34606
    2,4-Dinitrophenol            34616
    2-Methyl-4,6-dinitrophenol   34657
    2-Nitrophenol                34591
    4-Nitrophenol                34646
    Pentachlorophenol            39094
    Phenol                       34694
    2,4,6-Trichlorophenol        34621

    1.2   This  method  is applicable to the determination of these
         compounds in municipal  and industrial discharges.  It
         is designed to be used  to meet the monitoring
         requirements of the National Pollutant Discharge
         Elimination System  (NPDES).  As such, it presupposes a
                       nr-'K

-------
         high  expectation  of  finding  the  specific compounds  of
         interest.   If  the user  is  attempting  to  screen  samples
         for any or all  of the compounds  above, he must  develop
         independent protocols for  the  verification  of  identity.
    1.3  The sensitivity of this method is  usually dependent
         upon  the level  of interferences  rather than
         instrumental  limitations.  The limits of detection
         listed in  Table I represent  sensitivfties that  can  be
         achieved in wastewaters with a flame  ionization
         detector in the absence of interferences.   If  the
         derivatization  cleanup  is  required, the  sensitivity of
         the method is  10  ug/1.  This concentration  represents
         the minimum amount proven  to date  to  give reproducible
         and linear response  during derivatization.
    1.4  This method is  recommended for use only by  experienced
         residue analysts  or  under  the  close supervision  of  such
         qualified  persons.
2.   Summary of Method
    2.1  A 1-liter  sample  of  wastewater is  acidified  and
         extracted  with  methylene chloride  using  separatory
         funnel techniques.  The extract  is dried and
         concentrated to a volume of  10 ml  or  less.   Flame
         ionization gas chromatographic conditions are  described
         which allow for the  measurement  of the compounds in the
         extract.
    2.2  The method also provides for the preparation of
                       •nr-47

-------
         pentafluorobenzylbromide derivatives for electron
         capture gas chromatography with additional cleanup
         procedures to aid the analyst in the elimination of
         interferences.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample
         processing hardware may yield discrete artifacts and/or
         elevated baselines causing misinterpretation of gas
         chromatograms.  All of these materials must be
         demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.
         Specific selection of reagents and purification of
         solvents by distillation in all-glass systems may be
         required.
    3.2  Interferences coextracted from the samples will vary
         considerably from source to source, depending upon the
         diversity of the industrial complex or municipality
         being sampled.  While general cleanup techniques are
         provided as part of this method, unique samples may
         require additional cleanup approaches to achieve the
         sensitivities stated in Table I.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1  Grab sample bottle - amber glass, 1-liter or
                1-quart volume.  French or Boston Round design
                is recommended.  The container must be washed


-------
            and solvent rinsed before use to minimize
            interferences.
     4.1.2  Bottle caps - Threaded to screw on to the sample
            bottles.  Caps must be lined with Teflon.
     4.1.3  Compositing equipment - Automatic or manual
            compositing system.  Must incorporate glass
            sample containers for the collection of a
            minimum of 250 ml.  Sample containers must be
            kept refrigerated during sampling.  No tygon or
            rubber tubing may be used in this system.
4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
4.3  Drying column - 20 mm ID Pyrex chromatograhic column
     with coarse frit.
4.4  Kuderna-Danish (K-D) Apparatus
     4.4.1  Concentrator tube - 10 ml, graduated (Kontes
            K-570050-1025 or equivalent).  Calibration must
            be checked.  Ground glass stopper (size 19/22
            joint) is used to prevent evaporation of
            extracts.
     4.4.2  Evaporative flask - 500 ml (Kontes K-57001-0500
            or equivalent).  Attach to concentrator tube
            with springs.  (Kontes K-662750-0012).
     4.4.3  Snyder column - three-ball macro  (Kontes
            K-503000-0121 or equivalent).
     4.4.4  Snyder column - two-ball micro (Kontes
            K-569001-0219 or equivalent).

                        ft

-------
         4.4.5  Boiling chips - solvent extracted, approximately
                10/40 mesh.
    4.5  Water bath - Heated, with concentric ring cover,
         capable of temperature control (+ 2°C).  The bath
         should be used in a hood.
    4.6  Gas chromatograph - Analytical system complete with gas
         chromatograph suitable for on-column injection and all
         required accessories including flame ionization and
         electron capture detector, column supplies, recorder,
         gases, syringes.  A data system for measuring peak
         areas is recommended.
    4.7  Chromatographic column - 10 mm ID by 100 mm length,
         with Teflon stopcock.
    4.8  Reaction vial.
5.  Reagents
    5.1  Preservatives:
         5.1.1  Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2  Sulfuric acid - (1+1) Mix equal  volumes of cone.
                H2S04 (ACS) with distilled water.
         5.1.3  Sodium thiosulfate - (ACS) Granular.
    5.2  Methylene chloride, acetone, 2-propanol, hexane,
         toluene - Pesticide quality or equivalent.
    5.3  Sodium sulfate - (ACS) Granular, anhydrous  (purified by
         heating at 400°c for 4 hrs. in a shallow tray).
    5.4  Stock standards - Prepare stock standard solutions at a
         concentration of 1.00 ug/ul by dissolving 0.100 grams
                            so

-------
     of assayed reference material in pesticide quality
     2-propanol and diluting to volume in a 100 ml ground
     glass stoppered volumetric flask.  The stock solution
     is transferred to ground glass stoppered reagent
     bottles, stored in a refrigerator, and checked
     frequently for signs of degradation or evaporation,
     especially just prior to preparing working standards
     from them.
5.5  Sulfuric acid - (ACS) 1 N in distilled water.
5.6 ' Potassium carbonate - (ACS) powdered.
5.7  Pentafluorobenzyl bromide (a-Bromopentaf luorotoluene) -
     97% minimum purity.
5.8  1,4,7,10,13,16 - Hexaoxacyclooctadecane (18 crown 6) -
     98% minimum purity.
5.9  Derivatization reagent - Add 1 ml pentafluorobenzyl
     bromide and 1 gram 18 crown 6 to a 50 ml volumetric
     flask and dilute to volume with 2-propanol.  Prepare
     fresh weekly.
5.10 Silica gel - (ACS) 100/200 mesh, grade 923; activated
     at 130°C and stored in a desiccator.
Calibration
6.1  Prepare calibration standards for the flame ionization
     detector that contain the compounds of interest, either
     singly or mixed together.  The standards should be
     prepared at concentrations covering two or more orders
     of magnitude that will completely bracket the working
                     .TIT

-------
         range of the chromatographic system.  If the
         sensitivity of the detection system can be calculated
         from Table I as 100 ug/1 in the final extract, for
         example, prepare standards at 10 ug/1, 50 ug/1, 100
         ug/1, 500 ug/1, etc. so that injections of 1-5 ul of
         each calibration standard will define the linearity of
         the detector in the working range.
    6.2  Assemble the necessary gas chromatogrphic apparatus and
         establish operating parameters equivalent to those
         indicated in Table I.  By injecting calibration
         standards, establish the sensitivity limit of* the
         detector and the linear range of the analytical system
         for each compound.
    6.3  Before using the derivatization cleanup procedure, the
         analyst must process a series of calibration standards
         through the procedure to validate the precision of the
         derivatization and the absence of interferences from
         the reagents.
7.  Quality Control
    7.1  Before processing any samples, the analyst should
         demonstrate through the analysis of a distilled water
         method blank, that all glassware and reagents are
         interference-free.  Each time a set of samples is
         extracted or there is a -nange in reagents, a method
         blank should be process   = - a safeguard against
         chronic laboratory cc~      "ion.
                              .52

-------
    7.2  Standard quality assurance practices should be used
         with this method.  Field replicates should be collected
         to validate the precision of the sampling technique.
         Laboratory replicates should be analyzed to validate
         the precision of the analysis.  Fortified samples
         should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification
         of a peak on the chromatogram, confirmatory techniques
         such as mass spectroscopy should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.
         Conventional sampling practices should be followed,
         except that the bottle must not be prewashed with
         sample before collection.  Composite samples  should be
         collected in refrigerated glass containers in
         accordance with the requirements of the program.
         Automatic sampling equipment must be free of tygon and
         other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time
         of collection until extraction.  At the sampling
         location fill the glass container with sample.  Add 35
         mg of sodium thiosulfate per part per million free
         chlorine per liter.  Adjust the sample pH to
         approximately 2, as measured by pH paper, using
         appropriate sulfuric acid solution or 10 N sodium
         hydroxide.  Record the volume of acid used on the

-------
         sample identification tag so the sample volume can be
         corrected later.
    8.3  All samples must be extracted within 7 days and
         completely analyzed within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample
         bottle for later determination of sample volume.  Pour
         the entire sample into a two-liter separatory funnel.
         Adjust the sample pH to 12 with sodium hydroxide.
    9.2  Add 60 ml methylene chloride to the sample bottle,
         seal, and shake 30 seconds to rinse the inner walls.
         Transfer the solvent into the separatory funnel, and
         extract the sample by shaking the funnel for one minute
         with periodic venting to release vapor pressure.  Allow
         the organic layer to separate from the water phase for
         a minimum of ten minutes.  If the emulsion interface
         between layers is more than one-third the size of the
         solvent layer, the analyst must employ mechanical
         techniques to complete the phase separation.  The
         optimum technique depends upon the sample, but may
         include stirring, filtration of the emulsion through
         glass wool, or centrifugation.  Discard the methylene
         chloride layer, and wash the sarrple with an additional
         two 60 ml portions of methylene chloride in similar
         fashion.
    9.3  Adjust the aqueous layer to a p- sf 1-2 with sulfuric
         acid.

-------
9.4  Add 60 ml of methylene chloride to the sample and shake
     for two minutes.  Allow the solvent to separate from
     the sample and collect the methylene chloride in a 250
     ml Erlenmeyer flask.
9.5  Add a second 60 ml volume of methylene chloride to the
     sample bottle and complete the extraction procedure a
     second time, combining the extracts in the Erlenmeyer
     flask.
9.6  Perform a third extraction in the same manner.  Pour
     the combined extract through a drying column containing
     3-4 inches of anhydrous sodium sulfate, and collect it
     in a 500-ml Kuderna-Danish (K-D) flask equipped with a
     10 ml concentrator tube.  Rinse the Erlenmeyer flask
     and column with 20-30 ml methylene chloride to complete
     the quantitative transfer.
9.7  Add 1-2 clean boiling chips to the flask and attach a
     three-ball Snyder column.  Prewet the Snyder column by
     adding about 1 ml methylene chloride to the top.  Place
     the K-D apparatus on a hot water bath (60-65°C) so
     that the concentrator tube is partially immersed in the
     hot water, and the entire lower rounded surface of the
     flask is bathed in vapor.  Adjust the vertical position
     of the apparatus and the water temperature as required
     to complete the concentration in 15-20 minutes.  At the
     proper rate of distillation the balls of the column

                   1JL-   ^^

-------
     will actively chatter but the chambers will not flood.
     When the apparent volume of liquid reaches 1 ml, remove
     the K-D apparatus and allow it to drain for at least 10
     minutes while cooling.
9.8  Increase the temperature of the hot water bath to
     95-100°C.  Remove the Snyder column and rinse the
     flask and its lower joint into the concentrator tube
     with 1-2 ml of 2-propanol.  A 5-ml syringe is
     recommended for this operation.  Attach a micro-Snyder
     column to the concentrator tube and prewet the column
     by adding about 0.5 ml 2-propanol to the top.   Place
     the micro-K-D apparatus on the water bath so that the
     concentrator tube is partially immersed in the hot
     water.  Adjust the vertical position of the apparatus
     and the water temperature as required to complete
     concentration in 5-10 minutes.  At the proper rate of
     distillation, the balls of the column will actively
     chatter but the chambers will not flood.  When the
     apparent volume of the liquid reaches 2.5 ml, remove
     the K-D apparatus and allow it to drain for at least 10
     minutes while cooling.  Add an additional 2 ml of
     2-propanol  through the top of the micro-Snyder column
     and resume  concentrating as before.  When the apparent
     volume of liquid reaches 0.5 ml, remove the K-D
     apparatus and allow it to drain for at least 1C vnutes
     while cooling.  Remove the micro-Snyder colurn-

-------
         rinse its lower  joint  into the concentrator tube with a
         minimum of 2-propanol.   Adjust the extract volume to
         1.0 ml.  Stopper the concentrator tube and store in
         refrigerator,  if further processing will  not be
         performed immediately.   If the sample extract requires
         no further cleanup,  proceed with flame ionization gas
         chromatographic  analysis.   If the sample  requires
         cleanup, proceed to  Section 11.
    9.9  Determine the  original  sample volume by refilling the
         sample bottle  to the mark  and transferring the liquid
         to a 1000 ml graduated  cylinder.  After correction for
         sulfuric acid  preservative, record the sample volume to
         the nearest 5  ml.
10.  Gas Chromatography  -  Flame  Ionization Detector
    10.1 Table I summarizes some recommended gas chromatographic
         column materials and operating conditions for the
         instrument.  Included  in this table are estimated
         retention times  and  sensitivities that should be
         achieved by this method.  An example of the separation
         achieved by one  of these columns is shown in Figure 1.
         Calibrate the  gas chromatographic system  daily with a
         minimum of three injections of calibration standards.
    10.2 Inject 2-5 ul  of the sample extract using the
         solvent-flush  technique.  Smaller (1.0 ul) volumes can
         be injected if automatic devices are employed.  Record
         the volume injected to the nearest 0.05 ul, and the
                     Iff- -

-------
         resulting peak size, 1n area units.
    10.3 If the peak area exceeds the linear range of the
         system, dilute the extract and reanalyze.
    10.4 If the peak area measurement 1s prevented by the
         presence of Interferences, the phenols must be
         derlvatlzed and analyzed by electron capture gas
         chromatography.
11. Derivat1zation and Electron Capture Gas Chromatography
    11.1 P1pet a 1.0 ml aliquot of the 2-propanol solution of
         standard or sample extract Into a glass reaction vial.
         Add 1.0ml der1vat1zat1on reagent.  This 1s a
         sufficient anount of reagent to derlvatlze a solution
         whose total phenalic content does not exceed 0.3 mg/ml.
    11.2 Add about 3 mg of potassium carbonate to the solution
         and shake gently.
    11.3 Cap the mixture and heat 1t for 4 hours at SO°C 1n a
         hot water bath.
    11.4 Remove the solution from the hot water bath and allow
         1t to cool
    11.5 Add 10 ml hexane to the reaction vial and shake
         vigorously for one nrinute.  Add 3.0 ml of distilled,
         delonlzed water to the reaction vial and shake for two
         minutes.
    11.6 Decant organic layer into a concentrator tube and cap
         with a glass stopper.
    11.7 Pack a 10 mm ID chromatographic column with 4.0 grams
                             s-8

-------
     of activated silica gel.  After settling the silica gel
     by tapping the column, add about two grams of anhydrous
     sodium sulfate to the top.
11.8 Pre-elute the column with 6 ml hexane.  Discard the
     eluate and just prior, to exposure of the sulfate layer
     to air, pipet onto the column 2.0 ml of the hexane
     solution (11.6) that contains the derivatized sample or
     standard.  Elute the column with 10.0 ml of hexane
     (Fraction 1) and discard this fraction.  Elute the
     column, in order, with:  10.0 ml 15% toluene in hexane
     (Fraction 2); 10.0 ml 40% toluene in hexane (Fraction
     3); 10.0 ml 75% toluene in hexane (Fraction 4); and
     10.0 ml 15% 2-propanol in toluene (Fraction 5).
     Elution patterns for the phenolic derivatives are shown
     in Table II.  Fractions may be combined as desired,
     depending upon the specific phenols of  interest or
     level of interferences.
11.9 Analyze the fractions by electron capture gas
     chromatography.  Table II summarizes some recommended
     gas chromatographic column materials and operating
     conditions for the instrument.  Included in this table
     are estimated retention times that  should be achieved
     by this method.  Examples of the separation achieved by
     this  column  is shown  in Figure 2.   Calibrate the system
     daily with a minimum  of three aliquots  of calibration
     standards, containing each of the phenols of interest

-------
         that are derlvatlzed according to the procedure.
    11.10 Inject 2-5 ul of the column fractions using the
          solvent-flush technique.  Smaller (1.0 ul} volumes can
          be Injected 1f automatic devices are employed.  Record
          the volume Injected to the nearest 0.05 ul, and the
          resulting peak size, In area units.  If the peak area
          exceeds the linear range of the system, dilute the
          extract and reanalyze.
12. Calculations
    12.1  Determine the concentration of Individual compounds
          measured by the flame 1on1zat1on procedure (without
          der1vat1zat1on) according to the formula:
         Concentration, ug/1 « JA) (B) (Vt)
         where A » Calibration factor for chromatographlc
                   system, 1n nanograms material per area unit.
           B » Peak size in Injection of sample extract, 1n area
               units
           Vj » volume of extract Injected (ul)
           Vt » Volume of total extract (ul)
           V  » Volume of water extracted (ml)
    12.2  Determine the concentration of Individual compounds
          measured by the der1vat1zat1on and electron capture
          procedure according to the following procedure:
         12.2.1  From the concentration of the calibration
                 standards that were derlvatlzed with the

-------
        samples,  calculate the amounts, 1n nano grains,
        of underivatized phenols that were added as
        2-propanol solution (11.1)    From the size of
        the Injection Into the electron capture gas
        chromatograph, determine the nano grams of
        material  (calculated as the underivatized
        phenol) Injected onto the column.  Compare the
        detector responses obtained to develop a
        calibration factor for the chromatographlc
        system, 1n nanograms of material per area unit.
12.2.2  Determine the concentration of Individual
        compounds according to the formula:
        Concentration, ug/1 » (A) (B) (Vt)(10) (D)
       where A » Calibration factor for chromatographic
                 system, in nanograms material per area
                 unit, calculated as underivatized
                 phenol .
             B * Peak size 1n Injection of sample
                 extract, in area units.
             V1 » Volume of eluate injected (ul)
             Vt » Total volume of column eluate (ul)
               •
             V  * Volume of water extracted (ml)
             C * Volume of hexane sample solution added
                 to cleanup column, in ml.
             0 * Total volume of 2-propanol extract
              JJL

-------
                          after concentration.
                      E = Volume of 2-propanol extract used for
                          derivatization.
    12.3 Report results in micrograms per liter without
         correction for recovery data.  When  duplicate and
         spiked samples are analyzed, all data obtained should
         be reported.
13. Accuracy and Precision
    13.1 The U.S. EPA Environmental Monitoring and Support
         Laboratory in Cincinnati is in the process of
         conducting an interlaboratory method study to determine
         the accuracy and precision of this test procedure.
                           BIBLIOGRAPHY

"Development and Application of Test Procedures for Specific
Organic Toxic Substances in Wastewaters.  Category 3 -
Chlorinated Hydrocarbons and Category 8 - Phenols."  Report for
EPA Contract 68-03-2625 (In preparation).

-------
                             TABLE I
          Flame lonization Gas Chromatography of Phenols
Compound

2-Chlorophenol
2-Nitrophenol
Phenol
2,4-Dimethylphenol
2,4-Dichlorophenol
2,4,6-Trichlorophenol
4-Ch1oro-3-methyIpheno1
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
Pentachlorophenol
4-Nitrophenol
Retention Time
 Detection
Limit (ug/L)
1.70
2.00
3.01
4.03
4.30
6.05
7.50
10.00
10.24
12.42
24.25
2.0
2.5
1.4
1.7
2.1
5.0
8.3
7.0
10.0
10.0
10.0
Column conditions:  Supelcoport 80/100 mesh coated with 1%
SP-1240 DA in 6 ft long x 2 mm ID glass column with nitrogen
carrier gas at 30 ml/min flow rate.  Column temperature is
80°C at injection, programmed immediately at 8°C/min to
150°c final temperature.
                             - 63

-------
Detection limit is calculated from the minimum detectable GC
response being equal to five times the GC background noise,
assuming a 10 ml final extract volume of the 1 liter sample
extract, and assuming a GC injection of 5 micro!iters.

-------
                             TABLE II

               Electron Capture Gas Chromatography
                       of PFBB Derivatives

                          Retention     Recovery(%) by Fraction*
Parent Compound           Time (min)    J	2    3	4    5

2-chlorophenol                 3.3      -          90   <1
2-nitrophenol                  9.1      -     -          9   90
Phenol                         1.8      -          90   10
2,4-Dimethyphenol              2.9      -          95    7
2,4-Dichlorophenol             5.8      -          95   <1
2,4,6-Trichlorophenol          7.0           50    50
4-Chloro-2-methyphenol         4.8      -          84   14
Pentachlorophenol             28.8           75    20
4-Nitrophenol                 14.0      -     -     -   <1   90
(2,4-Dinitrophenol)           46.9**
(2-Methyl-4,6-dinitrophenol)  36.6**

Column conditions:  Chromosorb W-AW-DMCS 80/100 mesh coated with
5% OV-17 packed in a 1.8 m long x 2.0 mm ID glass column with 5%
methane/95% argon carrier gas at 30 ml/min flow r»te.  Column
temperature is 200°C.

-------
*From:  "Development and Application of Test Procedures for
Specific Organic Toxic Substances in Wastewaters.  Categories
3-Chlorinated Hydrocarbons and Category 8-Phenols."

**Retention times included for qualitative information only.
The lack of accuracy and precision of the derivatization
reaction precludes the use of this approach for quantitative
purposes.
                           'til

-------
      o

      UJ
  o-   I
  O   £

  q_i s    COLUMN:  1% SP-1240DA ON SUPELCOPORT

  ~'~ 5    PROGRAM: 80°C.-0 MINUTES 8° /MINUTE TO 15Q°C.
           DETECTOR:  FLAME IONIZATION
                                        o

                                        •2.
                                        LU
                                        O
                                        SS
                                       /
             8     12     16     20    2A


                RETENTION TIME-MINUTES
28
Figure 1.  Gas chromatogram of ph*
         IT

-------
                    COLUMN: 5% OV-17 ON CHROMOSORB W-AW
                    TEMPERATURE: 200°C.
                    DETECTOR:  ELECTRON  CAPTURE
                                                a.
                                                O
                                                cc
                                                O
              8      12     16     20     24

                 RETENTION TIME-MINUTES
28
32
Figure 2. Gas chromatogram of PFB derivatives of phenols
                     71!

-------
                                 BENZIDINES
                                 METHOD 605

1.   Scope and Application
    1.1   This method covers the determination of benzidine and selected
         derivatives.   The following parameters may be determined by this
         method:
                 Parameter                     STORET No.
              Benzidine                          39120
              3,3'-Dichlorobenzidine             34631
    1.2   This method is applicable to the measurement of these compounds
         in municipal  and industrial discharges.  It is designed to be
         used to  meet  the monitoring requirements of the National
         Pollutant Discharge Elimination System (NPDES).  As such, it
         presupposes a high expectation of finding the specific compounds
         of interest.   If the user is attempting to screen samples for any
         or all of the compounds above, he must develop independent
         protocols for the verification of identifications implied with
         the use  of these techniques.
    1.3   This sensitivity of this method is usually dependent upon the
         level of interferences rather than instrumental limitations.  The
         limits of detection listed in Table I represent sensitivities
         that can be achieved in wastewaters in the absence of inter-
         ferences.
    1.4   This method is recommended for use only by experienced residue
         analysts or under the close supervision of such qualified persons.

                              -  lol

-------
2.  Summary of Method
    2.1  Benzidine and 3,3-dichlorobenzidine (DCB) are extracted from the
         sample at pH 7-8 using chloroform.  The extract is then back
         extracted into acid, re-extracted into chloroform at neutral pH,
         and concentrated.  The benzidines are determined in the final
         extract using high performance liquid chromatography (HPLC) with
         electrochemical detection.
    2.2  If interferences are encountered with the measurement of benzi-
         dine, the method provides additional detector settings to
         increase the selectivity of the analytical system.
3.  Interferences
    3.1  Although the detection system is highly selective, solvents,
         reagents, glassware, and other sample processing hardware may
         yield discrete artifacts and/or elevated baselines causing
         misinterpretation of chromatograms.   All of these materials must
         be demonstrated to be free from interferences under the
         conditions of the analysis.  Specific selection of reagents and
         purification of solvents by distillation in all-glass systems may
         be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the
         industrial complex or municipality being sampled.  While general
         cleanup techniques are provided as part of this method,  unique
         samples may require additional cleanup approaches to achieve the
         sensitivities stated in Table I.
    3.3  Some dye plant effluents contain large amounts of components with

-------
         retention times in the vicinity of benzidine.  In these cases,-it
         has been found useful to reduce the electrode potential in order
         to eliminate the interferences but still detect the benzidine.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1  Grab sample bottle - amber glass, 1-pint or 1-quart
                volume. Quart bottles should be only half filled in the
                field.  French or Boston Round design is recommended.  The
                container should be washed and solvent rinsed before use
                to minimize interferences.
         4.1.2  Bottle caps - Threaded to screw on to the sample bottles.
                Caps must be lined with Teflon.  Foil may be substituted
                if the sample is not corrosive.
         4.1.3  Compositing equipment - Automatic or manual compositing
                system.  Must incorporate glass sample containers for the
                collection of a minimum of 250 ml.  Sample containers must
                be kept refrigerated during sampling.  No tygon or rubber
                tubing or fittings may be used in the system.
    4.2  Separatory funnel -  1000 ml and 250 ml, with Teflon stopcock.
    4.3  Water bath - Heated, with concentric ring cover, capable of
         temperature  control within 2°C.  The bath should be used in a
         hood.
    4.4  Rotating evaporator.
    4.5  Flasks -  round bottom, 100 ml, with 24/40 joints.
    4.6  Centrifuge tubes  - conical, screw capped, graduated, with Teflon
         lined caps.

-------
    4.7  Pipettes - Pasteur, with bulbs.
    4.8  High performance liquid chromatograph - Analytical system
         complete with column supplies, recorder, syringes, and the
         following components:
         4.8.1  Solvent delivery system - with pulse damper (Altex 110A,
                or equivalent).
         4.8.2  Injection valve Waters Model U6K (or equivalent).
         4.8.3  Electrochemical detector Bioanalytical Systems LC-2A with
                glassy carbon electrode, (or equivalent).
         4.8.4  Electrode polishing kit Princeton Applied Research Model
                9320 (or equivalent).
5.  Reagents
    5.1  Preservatives:
         5.1.1  Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2  Sulfuric acid - (ACS) Mix equal volumes of c'onc. H-S04
                with distilled water.
         5.1.3  Sodium thiosulfate - (ACS) granular.
    5.2  Sulfuric acid (1M) - (ACS) 58 ml/liter in distilled water.
    5.3  Sodium hydroxide (20%) - (ACS) 20 grams/100 ml in distilled water.
    5.4  Sodium hydroxide (1M) - (ACS) 40 grams/liter in distilled water.
    5.5  Sodium tribasic phosphate (0.4M) - (ACS) 160 grams
         Na3P04(12H20) in 1 liter of distilled water.
    5.6  Acetate buffer 0.1M, pH 4.7; 5.8 mis glacial acetic acid (ACS)
         and 13.6 grams of sodium acetate trihydrate (ACS) per liter in
         distilled water.
    5.7 ' Acetonitrile - Pesticide quality or equivalent.

                      W~   7O

-------
    5.8  Methyl  alcohol  -  Pesticide quality or equivalent.
    5.9  Chloroform (preserved with 1% ETOH) - Burdick and Jackson (or
         equivalent).
    5.10 Water - Purified  (e.g.  from Millipore RO-4 System or equivalent).
    5.11 HPLC mobile phase - Place 1 liter of filtered (through Millipore
         type FH filter, or equivalent) acetonitrile and 1 liter of
         filtered (through Millipore type GS filter, or equivalent)
         acetate buffer  in a narrow mouth, glass, one gallon jug and mix
         thoroughly.  Prepare fresh weekly.
    5.12 Stock standards - Prepare stock standard solutions at a concen-
         tration of 0.100  ug/ul  by dissolving 0.0100 grams of assayed
         reference material in pesticide quality methyl alcohol and
         diluting to volume in a 100 ml ground glass stoppered volumetric
         flask.  The stock solution is transfered to ground glass
         stoppered reagent bottles, stored in a refrigerator, and checked
         frequently for  signs of degradation or evaporation, especially
         just prior to preparing working standards from them.
6.  Calibration
    6.1  Prepare calibration standards that contain the compounds of
         interest, either  singly or mixed together.  The standards should
         be prepared from  the stock standards at the following concen-
         trations that will bracket the working range of the
         chromatographic system:

-------
     Solution Cone, (nq/ul)       Sensitivity fnA Full Scale)
             0.05                             10
             0.10                             10
             0.50                             50
             1.0                             100
             5.0                             500
6.2  Assemble the necessary liquid chromatographic apparatus and
     establish operating parameters equivalent to those Indicated 1n
     Table I.
6.3  A constant Injection volume of 25 microliters should be employed
     for all subsequent measurements.
6.4  In order to determine the precision of the HPLC system, a series
     of 6 replicate Injections of a 1 ng/ul solution of benzldlne and
     3,3'HJlchlorobenz1d1ne (DCS) should be made on the 50nA full
     scale setting.  A precision of +43J should be achieved for the
     peak heights of both benzldlne and DCS.  This measurement should
     be made every few weeks or whenever Instrument related problems
     are apparent.  A sample chromatogram is shown 1n Figure 1.
6.5  Retention times should remain relatively constant (within +5S day
     to day) with benzldlne being  6.1 minutes and DCS-being  12.1
     minutes under the specified conditions.  These values should be
     checked daily when the calibration injections are made.
6.6  If serious loss of response occurs, it may be necessary to polish
     the surface of the carbon electrode (according to the instruc-
     tions supplied with the polishing kit).  In this case, it will be
     necessary to recalibrate the system.

-------
    6.7  When leaving the Instrument Idle, It is advisable to maintain  a
         flow of 0.1 inl/m1n of mobile phase through the HPLC column  in
         order to prolong column life.
7.  Quality Control
    7.1  Before processing any samples, the analyst should demonstrate
         through the analysis of a distilled water blank, that all
         glassware and reagents are interference-
         free.  Each time a set of samples is extracted or there is  a
         change in reagents, a method blank should be processed as a
         safeguard against laboratory contamination.
    7.2  Standard quality assurance practices should be used with this
         method.  Field replicates should be collected to validate the
         precision of the sanpling technique.  Laboratory replicates
         should be analyzed to validate the precision of the analysis.
         Fortified samples should be analyzed to validate the accuracy  of
         the analysis.  Where doubt exists over the identification of a
         peak on the chromatogram, confirmatory techniques such as mass
         spectroscopy should be used.
8.  Sample Collection. Preservation, and Handling
    3.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle
         must not be prewashed with sample before collection.  Composite
         samples should be collected in refrigerated glass containers  in
         accordance with the requirements of the program.  Automatic
         sampling equipment must be free  of tygon  and other potential
                           n  7
                            '

-------
         sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of
         collection to extraction.  Benzidine and dichlorobenzidine are
         easily oxidized by materials such as free chlorine.  For
         chlorinated wastes, immediately add 35 mg sodium thiosulfate per
         part per million of free chlorine per liter.
    8.3  If 1,2-diphenylhydrazine is likely to be present, adjust the pH
         of the sample to 4+0.2 units to prevent rearrangement to
         benzidine.  Otherwise, if the samples will not be extracted
         within 48 hours of collection, the sample pH should be adjusted
         to 2-3 with sodium hydroxide or sulfuric acid.
    8.4  All samples must be extracted within 7 days and completely
         analyzed within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for
         later determination of sample volume.  Quantitatively pour the
         entire sample into a one-liter separatory funnel.  Check the pH
         of the sample with wide-range pH paper and adjust to within the
         range of 6.5 - 7.5 with sodium hydroxide or sulfuric acid.
    9.2  Add 100 ml chloroform to the sample bottle, seal, and shake 30
         seconds to rinse the inner walls.  Transfer the solvent into the
         separatory funnel, and extract the sample by shaking the funnel
         for two minutes.  Allow the organic layer to separate from the
         sample a minimum of ten minutes, then collect the chloroform
         extract in a 250-ml separatory funnel.
    9.3  Add 50 ml chloroform to the sample bottle and complete the

-------
     extraction procedure a second time, combining the extracts  in  the
     separatory funnel.  Perform a third extraction in the same  manner
     with an additional 50 ml volume.
9.4  Discard any aqueous layer from the 250 ml funnel containing the
     combined organic layers.  Add 25 ml of 1 M sulfuric acid and
     extract the sample by shaking the funnel for two minutes.
     Transfer the aqueous layer to a 250 ml beaker.  Extract with two
     additional 25 ml portions of 1 M sulfuric acid and combine  the
     acid extracts in the beaker.
9.5  Place a stirbar in the 250 ml beaker and stir the acid extract
     while carefully adding 5 ml of a .4M sodium tribasic phosphate.
     With the aid of a pH meter neutralize the extract to pH 6-7 by
     dropwise addition of 20 percent NaOH while stirring the solution
     vigorously.  Approximately 25-30 ml of 20 percent NaOH will be
     required and it should be added over at  least a 2-minute period.
     Do not allow the sample pH ever to exceed pH 8.
9.6  Pour the neutralized extract into a 250 ml separatory funnel.
     Add 30 ml of chloroform and shake the funnel for 2 minutes.
     Allow phases to separate, and transfer the organic layer in a
     second 250 ml separatory funnel.
9.7  Extract the aqueous layer with two additional 20 ml aliquots of
     chloroform as before.  Combine the extracts in the 250 ml
     separatory funnel.
9.8  Add 20 ml of distilled water to the combined organic  layers,
     shake for 30 seconds, and discard aqueous layer.
9.9  Transfer the organic extract into a 100 ml round bottom flask.

                      T/T- T5"

-------
         Add 20 ml of methyl alcohol and concentrate to  5 ml on a
         rotating evaporator at 35°C.
    9.10 Using a 9-inch Pasteur pipette, transfer the extract to a 15 ml
         conical screw-capped centrifuge tube.  Rinse the flask, including
         the entire side wall, with two ml portions of methyl alcohol and
         combine with the original extract.
    9.11 Carefully concentrate the extract to 0.5 ml using a gentle stream
         of nitrogen and a 30°C water bath, dilute to 2 ml with methyl
         alcohol, reconcentrate to 1 ml, and dilute to 5 ml with acetate
         buffer.  Mix extract thoroughly.  Stopper the ampul and store
         refrigerated if further processing will not be performed
         immediately.
    9.12 Determine the original sample volume by refilling the sample
         bottle to the mark and transferring the liquid to a 1000 ml
         graduated cylinder.  Record the sample volume to the nearest 5 ml.
10.  Liquid Chromatography fHPLC)
    10.1 Table I summarizes the recommended liquid chromatographic column
         material and operating conditions for the instrument.  Included
         in this table are estimated retention times and sensitivities
         that should be achieved by this method.  An example of the
         separation achieved by this column is shown in Figure 1.
         Calibrate the system daily with a minimum of three injections of
         calibration standards.
    10.2 Inject 25 ul of the sample extract.   If the peak area exceeds the
         linear range of the system, dilute the extract and reanalyze.
    10.3 If the peak area measurement for benzidine is prevented by the

-------
         presence of interferences, reduce the electrode potential to 0.7
         V and reanalyze.
11.  Calculations
    11.1  Determine the concentration of individual compounds according to
         the formula:
         Concentration, ug/1 = (A)  (B)  (Vt)
                                 (V^  (Vs)
         where A = Calibration factor for chromatographic
                   system, in nanograms material per unit peak area.
               B = Peak size in injection of sample extract, in
                   peak area units
               V.. = volume of extract injected (ul)
               Vt = Volume of total extract (ul)
               V  = Volume of water extracted (ml)
    11.2 Report results in micrograms per liter without correction for
         recovery data.  When duplicate and spiked samples are analyzed,
         all data obtained should be reported.
12.  Accuracy and Precision
    12.1 The U.S. EPA Environmental Monitoring and Support Laboratory in
         Cincinnati is in the process of conducting an interlaboratory
         method study to determine the accuracy and -    -  'on of this test
         procedure.

-------
                           BIBLIOGRAPHY

"Development and Application of Test  Procedures  for  Specific
Organic Toxic Substances in Wastewaters Category 7 -
Benzidines," Report for EPA Contract  68-03-2624  (In  preparation)
                         71-

-------
                             TABLE I
               Liquid Chromatography of Benzidines

                             Retention             Detection
Compound                    Time (min.)           Limit  fug/1)

Benzidine                       6.1                    0.05
3,3'-dich1orobenzidine         12.1                    0.1
Column conditions:  Lichrosorb RP-2,  5 micron particle diameter,
    packed in a 25 cm x 4.6 mm ID stainless steel  column  with
    0.8 ml/min flow rate of mobile phase (50% acetonitrile:  50%
    0.1 M pH 4.7 acetate buffer).

 Detection limit is calculated from the minimum detectable
    response of the electrochemical detector being equal  to  five
    times the background noise, assuming a 5 ml final  extract
    volume of the 500 ml sample, and assuming an injection
    volume of 25 micro!iters.
                      sr-7?

-------
     uu


     Q

     INI
     Z
     LU
     03
            CO
            O
            ee
            O


            O
             PO
0    4    8   12


  RETEMTION TINIE-WINUTES


Figure 1. Liquid chromatogram of benzidines
nt

-------
                         PHTHALATE ESTERS
                            METHOD 606

1.  Scope and Application
    1.1  This method covers the determination of certain
         phthalate esters.  The following parameters may be
         determined by this method:
         Parameter                                    STORET No.
         Benzyl butyl phthalate                        34292
         B1s(2-ethy1hexy1) phthalate                   39100
         01-n-butyl phthalate                          34110
         01-n-octyl phthalate                          34596
         01ethyl phthalate                             34336
         Dimethyl phthalate                            34341
    1.2  This method 1s applicable to the determination of these
         compounds In municipal and industrial discharges.  It
         1s designed to be used to meet the monitoring require-
         ments of the National Pollutant Discharge Elimination
         System (NPOES).  As such, it presupposes a high expec-
         tation of finding the specific compounds of  interest.
         If the user is attempting to screen  samples  for any or
         all of the compounds above, he must  develop  independent
         protocols for the verification of identity.
    1.3  The sensitivity  of this method is usually dependent
         upon  the  level of interferences rather  than  instru-
         mental limitations.  The  limits of detection listed  in

-------
         Table I represent sensitivities that can be achieved in
         wastewaters in the absence of interferences.
    1.4  This method is recommended for use only by experienced
         residue analysts or under the close supervision of such
         qualified persons.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with
         methylene chloride using separatory funnel techniques.
         The extract is dried and concentrated to a volume of 10
         ml or less.  Chromatographic conditions are described
         which allow for the accurate measurement of the com-
         pounds in the extract.
    2.2  If interferences are encountered, the method provides
         selected general purpose cleanup procedures to aid the
         analyst in their elimination.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample pro-
         cessing hardware may yield discrete artifacts and/or
         elevated baselines causing misinterpretation of gas
         chromatograms.  All of these materials must be
         demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.
         Specific selection of reagents and purification of
         solvents by distillation in all-glass systems may be
         required.
    3.2  Interferences coextracted from the samples will vary

-------
         considerably from source to source, depending upon the
         diversity of the industrial complex or municipality
         being sampled.  While general cleanup techniques are
         provided as part of this method, unique samples may
         require additional  cleanup approaches to achieve the
         sensitivities stated in Table I.
    3.3  Phthalate esters contaminate many types of products
         commonly found in the laboratory.  The analyst must
         demonstrate that no phthalate residues contaminate the
         sample or solvent extract under the conditions of the
         analysis.  Of particular importance is the avoidance of
         plastics because phthalates are commonly used as
         plasticizers and are easily extracted from plastic
         materials.  Serious, phthalate contamination may result
         at any time if consistent quality control  is not
         practiced.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1     Grab sample bottle - amber glass, 1-liter or
                   1-quart volume.  French or Boston Round
                   design is recommended.  The container must be
                   washed and solvent rinsed before use to
                   minimize interferences.
         4.1.2     Bottle caps - Threaded to screw on to the
                   sample bottles.  Caps must be lined with
                   Teflon.  Foil may be substituted if sample is

-------
               not corrosive.
     4.1.3     Compositing equipment - Automatic or manual
               compositing system. Must incorporate glass
               sample containers for the collection of a
               minimum of 250 ml.  Sample containers must be
               kept refrigerated during sampling.  No tygon
               or rubber tubing may be used in the system.
4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
4.3  Drying column - 20 mm ID pyrex chromatographic column
     with coarse frit.
4.4  Kuderna-Danish (K-D) Apparatus
     4.4.1     Concentrator tube - 10 ml, graduated (Kontes
               K-570050-1025 or equivalent).  Calibration
               must be checked.  Ground glass stopper (size
               19/22 joint) is used to prevent evaporation
               of extracts.
     4.4.2     Evaporative flask - 500 ml (Kontes
               K-57001-0500 or equivalent).  Attach to
               concentrator tube with springs.  (Kontes
               K-662750-0012).
     4.4.3     Snyder column - three-ball macro (Kontes
               K503000-0121 or equivalent).
     4.4.4     Snyder column - two-ball micro (Kontes
               K-569001-0219 or equivalent).
     4.4.5     Boiling chips - solvent extracted, approxi-
               mately 10/40 mesh.

-------
    4.5  Water bath - Heated, with concentric ring cover,
         capable of temperature control (+ 2°C).  The bath
         should be used in a hood.
    4.6  Gas chromatograph - Analytical system complete with gas
         chromatograph suitable for on-column injection and all
         required accessories including electron capture or
         flame ionization detector, column supplies, recorder,
         gases, syringes.  A data system for measuring peak
         areas is recommended.
    4.7  Chromatography column - 300 mm long x 10 mm ID with
         coarse fritted disc at bottom and Teflon stopcock
         (Kontes K-420540-0213 or equivalent).
5.  Reagents
    5.1  Preservatives:
         5.1.1     Sodium hydroxide - (ACS) 10 N in distilled
                   water.
         5.1.2     Sulfuric acid - (ACS) Mix equal volumes of
                   cone. H2S04 with distilled water.
    5.2  Methylene chloride - Pesticide quality or equivalent.
    5.3  Sodium Sulfate - (ACS) Granular, anhydrous (purified by
         heating at 400°C for 4 hrs. in a shallow tray).
    5.4  Stock standards - Prepare stock standard solutions at a
         concentration of 1.00 ug/ul by dissolving 0.100 grams
         of assayed reference material in pesticide quality
         isooctane or other appropriate solvent and diluting to
         volume in a  100 ml ground glass stoppered volumetric
                        ITT

-------
         flask.   The stock solution is transferred to ground
         glass stoppered reagent bottles, stored in a refriger-
         ator, and checked frequently for signs of degradation
         or evaporation, especially just prior to preparing
         working standards from them.
    5.5  Diethyl Ether - Nanograde, redistilled in glass if
         necessary.
         5.5.1     Must be free of peroxides as indicated by EM
                   Quant test strips.  (Test strips are avail-
                   able from EM Laboratories, Inc., 500
                   Executive Blvd., Elmsford, N.Y.  10523.)
         5.5.2     Procedures recommended for removal of
                   peroxides are provided with the test strips.
                   After cleanup, 20 ml ethyl alcohol preserva-
                   tive must be added to each liter of ether.
    5.6  Florisil - PR grade (60/100 mesh); purchase activated
         at 1250°F and store in dark in glass container with
         ground glass stoppers or foil-lined screw caps.
    5.7  Alumina - Activity Super I, Neutral, W200 series, (ICN
         Life Sciences Group, No. 404583).
    5.8  Hexane - Pesticide quality.
6.  Calibration
    6.1  Prepare calibration standards that contain the com-
         pounds of interest, either singly or mixed together.
         The standards should be prepared at concentrations
         covering two or more orders of magnitude that will
                         72T-

-------
         completely bracket  the working range of the
         chromatographic  system.   If  the sensitivity of  the
         detection  system can  be calculated  from Table I as 100
         ug/1  in  the final extract, for example,  prepare stan-
         dards at 10 ug/1, 50  ug/1,  100 ug/1, 500 ug/1,  etc.  so
         that  injections  of  1-5 ul  of each calibration standard
         will  define the  linearity of the detector in the work-
         ing range.
    6.2  Assemble the necessary gas chromatographic apparatus
         and establish operating parameters  equivalent to those
         indicated  in Table  I.  By injecting calibration stan-
         dards, establish the  sensitivity limit of the detector
         and the linear range  of the  analytical system for each
         compound.
    6.3  Before using any cleanup procedure, the analyst must
         process a series of calibration standards through the
         procedure to validate elution patterns and the absence
         of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any samples, the  analyst should
         demonstrate through the analysis of a distilled water
         method blank, that  all glassware and reagents are
         interference-free.   Each time a set of samples is
         extracted or there  is a change in reagents, a method
         blank should be processed as a safeguard against
         chronic laboratory contamination.

                          JT-ffc

-------
    7.2  Standard quality assurance practices should be used
         with this method.  Field replicates should be collected
         to validate the precision of the sampling technique.
         Laboratory replicates should be analyzed to validate
         the precision of the analysis.  Fortified samples
         should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification
         of a peak on the chromatogram, confirmatory techniques
         such as mass spectroscopy should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.
         Conventional sampling practices should be followed,
         except that the bottle must not be prewashed with
         sample before collection.  Composite samples should be
         collected in refrigerated glass containers in accor-
         dance with the requirements of the program.  Automatic
         sampling equipment must be free of tygon and other
         potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time
         of collection until extraction.  Chemical preservatives
         should not be used in the field unless more than 24
         hours will elapse before delivery to the laboratory.
         If the samples will not be extracted within 48 hours of
         collection, the sample should be adjusted to a pH range
         of 6.0-8.0 with sodium' hydroxide or sulfuric acid.
    8.3  All samples must be extracted within 7 days and

                           777-

-------
         completely analyzed within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the  sample
         bottle for later determination of sample volume.  Pour
         the entire sample into a two-liter separatory funnel.
         Check the pH of the sample pH with wide-range paper and
         adjust to within the range of 5-9 with sodium hydroxide
         or sulfuric acid.
    9.2  Add 60 ml methylene chloride to the sample  bottle,
         seal, and shake 30 seconds to rinse the inner walls.
         Transfer the solvent into the separatory funnel, and
         extract the sample by shaking the funnel for two
         minutes with periodic venting to release vapor pres-
         sure.  Allow the organic layer to separate  from the
         water phase for a minimum of ten minutes.   If the emul-
         sion interface between layers is more than  one-third
         the size of the solvent layer, the analyst  must employ
         mechanical techniques to complete the phase separa-
         tion.  The optimum technique depends upon the sample,
         but may include stirring, filtration of the emulsion
         through glass wool, or centrifugation.  Collect the
         methylene chloride extract in a 250-ml Erlenmeyer flask,
    9.3  Add a second 60-ml volume of methylene chloride to the
         sample bottle and complete the extraction procedure a
         second time, combining the extracts in the  Erlenmeyer
         flask.

                          TTt--7

-------
9.4  Perform a third extraction in the same manner.  Pour
     the combined extract through a drying column containing
     3-4 inches of anhydrous sodium sulfate, and collect it
     in a 500-ml Kuderna-Danish (K-D) flask equipped with a
     10 ml concentrator tube.  Rinse the Erlenmeyer flask
     and column with 20-30 ml methylene chloride to complete
     the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a
     three-ball Snyder column.  Prewet the Snyder column by
     adding about 1 ml methylene chloride to the top.  Place
     the K-D apparatus on a hot water bath (60-65°C) so
     that the concentrator tube is partially immersed in the
     hot water, and the entire lower rounded surface of the
     flask is bathed in vapor.  Adjust the vertical position
     of the apparatus and the water temperature as required
     to complete the concentration in 15-20 minutes.  At the
     proper rate of distillation the balls of the column
     will actively chatter but the chambers will not flood.
     When the apparent volume of liquid reaches 1 ml, remove
     the K-D apparatus and allow it to drain for at least 10
     minutes while cooling.
9.6  Increase the temperature of the hot water bath to about
     80°C.  Momentarily remove the Snyder column, add 50
     ml of hexane and a new boiling chip and reattach the
     Snyder column.  Pour about 1 ml of hexane into the top
     of the Snyder column and concentrate the solvent

-------
         extract as before.   Elapsed time of concentration
         should be 5 to 10 minutes.   When the apparent volume of
         liquid reaches 1  ml,  remove the K-D apparatus and allow
         it  to drain at least  10 minutes while cooling.  Remove
         the Snyder column and rinse the flask and its lower
         joint into the concentrator tube with 1-2 ml  of hexane,
         and adjust the volume to 10 ml.  A 5-ml  syringe is
         recommended for this  operation.  Stopper the  concen-
         trator tube and store refrigerated if further process-
         ing will  not be performed immediately.  If the sample
         extract requires no further cleanup, proceed  with gas
         chromatographic analysis.  If the sample requires
         cleanup,  proceed to Section 10.
    9.7  Determine the original  sample volume by  refilling the
         sample bottle to the  mark and transferring the liquid
         to  a 1000 ml graduated  cylinder.  Record the  sample
         volume to the nearest 5 ml.
10.  Cleanup  and Separation
    10.1  If  the entire extract is to be cleaned up by  one of the
         following two procedures, it must be concentrated to
         about 2 ml .  To the concentrator tube in 9.6, add a
         clean boiling chip  and  attach a two-ball micro-Snyder
         column.  Prewet the column by adding about 0.5 ml
         hexane through the  top.  Place the K-D apparatus on a
         hot water bath (80°C) so that the concentrator tube
         is  partially immersed in the hot water.   Adjust the
                         is

-------
     vertical  position of the apparatus and the water
     temperature as required to complete the concentration
     in 5-10 minutes.   At the proper rate of distillation
     the balls of the  column will  actively chatter but the
     chambers  will  not flood.  When the apparent volume of
     liquid reaches about 0.5 ml,  remove the K-D apparatus
     and allow it to drain for at  least 10 minutes while
     cooling.   Remove  the micro-Snyder column and rinse its
     lower joint into  the concentrator tube with 0.2 ml of
     hexane.  Proceed  with one of  the following clean-up
     procedures.
10.2 Florisil  Column Cleanup for Phthalate Esters
     10.2.1 Place lOOg of Florisil into a 500 ml beaker and
            heat for approximately 16 hours at 400°C.
            After heating transfer to a 500 ml reagent
            bottle.  Tightly seal  and cool to room temper-
            ature.   When cool add  3 ml of distilled water
            which is free of phthalates and interferences.
            Mix thoroughly by shaking or rolling for 10
            minutes and let it stand for at least 2 hours.
            Keep the bottle sealed tightly.
     10.2.2 Place lOg  of this Florisil preparation into a
            10 mm ID chromatography column and tap the
            column to  settle the Florisil.  Add 1 cm of
            anhydrous  sodium sulfate to the top of the
            Florisil.
                   70

-------
     10.2.3  Preelute the column with 40 ml  of hexane.   Dis-
            card  this eluate and just prior to exposure of
            the sodium sulfate layer to the air transfer the
            2  ml  sample extract onto the column,  using an
            additional 2 ml  of hexane complete the transfer.
     10.2.4  Just  prior to exposure of the sodium  sulfate
            layer to the air add 40 ml hexane and continue
            the elution of the column. Discard this hexane
            eluate.
     10.2.5  Next  elute the phthalate esters with  100 ml of
            20% ethyl ether /80% hexane (V/V) into a 500 ml
            K-D flask equipped with a 10 ml concentrator
            tube.  Elute the column at a rate of  about 2 ml
            per minute for all fractions.  Concentrate the
            collected fraction by standard  K-D technique.
            No solvent exchange is necessary.  After
            concentration and cooling, adjust the volume of
            the cleaned up extract to 10 ml in the concen-
            trator tube and analyze by gas  chromatography.
10.3 Alumina Column Cleanup for Phthalate Esters
     10.3.1  Place 100 g of alumina into a 500 ml  beaker and
            heat for approximately 16 hours at 400°C.
            After heating transfer to a 500 ml reagent
            bottle.  Tightly seal and cool  to room
            temperature.  When cool add 3 ml of distilled
            water which is free from phthalates and
                               2-

-------
       interferences.   Mix thoroughly by shaking or
       rolling for 10  minutes and let it stand for at
       least 2 hours.   Keep the bottle sealed tightly.
10.3.2 Place lOg of this alumina preparation into a 10
       mm ID chromatography column and tap the column
       to settle the alumina.  Add 1 cm of anhydrous
       sodium sulfate  to the top of the alumina.
10.3.3 Preelute the column with 40 ml of hexane.  Dis-
       card this eluate and just prior to exposure of
       the sodium sulfate layer to the air, transfer
       the 2 ml sample extract onto the column, using
       an additional 2 ml of hexane to complete the
       transfer.
10.3.4 Just prior to exposure of the sodium sulfate
       layer to the air add 35 ml hexane and continue
       to elution of the column.  Discard this hexane
       eluate.
10.3.5 Next elute the  column with 140 ml of 20% ethyl
       ether/80% hexane (V/V) into a 500 ml K-D flask
       equipped with a 10 ml concentrator tube.  Elute
       the column at a rate of about 2 ml per minute
       for all fractions.  Concentrate the collected
       fraction by standard K-D technique.  No solvent
       exchange is necessary.  After concentration and
       cooling adjust  the volume of the cleaned up
       extract to 10 ml in the concentrator tube and
                     -13

-------
                analyze by gas chromatography.
11.  Gas Chromatography
    11.1 Table I summarizes some recommended gas chromatographic
         column materials and operating conditions for the
         instrument.  Included in this table are estimated
         retention times and sensitivities that should be
         achieved by this method.  Examples of the separations
         achieved by the primary column are shown in Figures 1
         and 2.  Calibrate the system daily with a minimum of
         three injections of calibration standards.
    11.2 Inject 2-5 ul of the sample extract using the
         solvent-flush technique.  Smaller (1.0 ul) volumes can
         be injected if automatic devices are employed.  Record
         the volume injected to the nearest 0.05 ul, and the
         resulting peak size, in area units.
    11.3 If the peak area exceeds the linear range of the
         system, dilute the extract and reanalyze.
    11.4 If the peak area measurement is prevented by the
         presence of interferences, further cleanup is required.
12. Calculations
    12.1 Determine the concentration of individual compounds
         according to the formula:
         Concentration, ug/1 = (A) (B) (Vt)
                                 (Vi)  (Vs)
         where A = Calibration factor for chromatographic
                   system, in nancgrams material per area unit.
                         \ll-~

-------
               B « Peak size in injection of sample extract, in
                   area units
               V. = volume of extract injected (ul)
               Vt = Volume of total extract (ul)
               V$ = Volume of water extracted (ml)
    12.2 Report results in micrograms per liter without correc-
         tion for recovery data.  When duplicate and spiked
         samples are analyzed, all data obtained should be
         reported.
13. Accuracy and Precision
    13.1 The U.S. EPA Environmental Monitoring and Support
         Laboratory in Cincinnati is in the process of conduct-
         ing a interlaboratory method study to determine the
         accuracy and precision of this test procedure.

                           BIBLIOGRAPHY
"Development and Application of Test Procedures for Specific
Organic Toxic Substances in Wastewaters.  Category
1-Phthalates."  Report for EPA Contract 68-03-2606 (In
preparation).
                         ¥•   9*"

-------
                             TABLE I
              GAS CHROMATOGRAPHY OF PHTHALATE ESTERS
                  Retention Time (min.)    Detection Limit (ug/1)
Compound	Column 1   Column 2	EC	FID
Dimethyl phthalate
Diethyl phthalate
Di-n-butyl phthalate
Butyl benzyl
phthalate
Bis(2-ethylhexyl)
phthalate
Di-n-octyl
phthalate
2.03
2.82
8.65
6.94*
8.92*
16.2*
0.95
1.27
3.50
5.11**
10.5**
18.0**
0.11
0.13
0.02
0.02
0.04
0.11
19
31
14
15
20
31
Column 1 conditions:  Supelcoport ToO/120 mesh coated with 1.5%
    SP-2250/1.95% SP-2401 packed in a 180 cm long x 4 mm ID
    glass column with carrier gas at 60 ml/min flow rate.
    Column temperature is 180°C except where * indicates
    220°C.  Under these conditions R.T. of Aldrin is 5.49 min.
    at 180°C and 1.84 min at 220°C.
Column 2 conditions:  Supelcoport 100/120 mesh coated with 3%
    OV-1 in a 180 cm long x 4 mm ID glass column with carrier
    gas at 60 ml/min flow rate.  Column temperature is 200°C
    except where ** indicates 220°C.  Under these conditions
    R.T. of Aldrin  is 3.18 min. at 200°C and 1.46 min. at
    220°C.
  Detection limit is calculated from the minimum detectable GC
    response being  equal to five times the GC background noise,
    assuming a  10 ml final volume of the 1 liter sample  extract,
    and assuming a  GC injection of 5 micro!iters.

-------
  I  COLUMN: 1.5% SP-2250 +
            1.95% SP-2401 ON SUPELCOPORT
    TEMPERATURE:  180°C.
    DETECTOR;  ELECTRON CAPTURE
     <
     _i
     <
     x
                 a.
                 oa
                 a
0   2   4   6   8  10  12
  RETENTION TIME-MINUTES

Figure 1.  Gas chromatogram of phthalates
               •a.

-------
COLUMN:  1.5% SP-2250 +

         1.95% SP-2401 ON SUPELCOPORT

TEMPERATURE: 180°C.

DETECTOR: ELECTRON CAPTURE
              X
              a.
           X  X
           a.  t—

           _a  ^

           *"  d


           LU  oa
           ca
                          X
                          a.
O

c
0      4      8     12     16      :


      RETENTION TIME-MINUTES



Figure 2. Gas  chromatogram c* phthalates

-------
                                 NITROSAMINES
                                  METHOD 607

1.   Scope and Application
    1.1   This method covers the determination of certain nitrosamines.  The
         following parameters may be determined by this method:
         Parameter                                         STORET No.
         N-nitrosodimethylamine                            34438
         N-nitrosodiphenylamine                            34433
         N-nitrosodi-n-propylamine                         34428
    1.2   This method is applicable to the determination of these compounds
         in municipal  and industrial discharges.  It is designed to be used
         to meet the monitoring requirements of the National  Pollutant
                                                   /
         Discharge Elimination System (NPDES).   As such, it presupposes a
         high expectation of finding the specific compounds of interest.  If
         the user is attempting to screen samples for any or  all of the
         compounds above, he must develop independent protocols for the
         verification  of identity.
    1.3   The sensitivity of this method is usually dependent  upon the level
         of interferences rather than instrumental limitations.  The limits
         of detection  listed in Table I represent sensitivities that can be
         achieved in wastewaters in the absence of interferences.
    1.4   This method is recommended for use only by experienced residue
         analysts or under the close supervision of such qualified persons.
    1.5   The analyst must understand that nitrosamines are known carcino-
         gens.   Utmost care must be exercised in the handling of materials

-------
         which are known or believed to contain nitrosanrines.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
         using separatory funnel techniques.  The extract is dried and
         concentrated to a volume of 10 ml or less.  Depending upon the
         nitrosamines being measured, a column cleanup procedure may be
         required.  Chromatographic conditions are described which allow for
         the accurate measurement of the compounds in the extract.
    2.2  If interferences are encountered, the method provides selected
         general purpose cleanup procedures to aid the analyst in their
         elimination.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
         in all-glass systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the indus-
         trial complex or municipality being sampled.  While general
         clean-up techniques are provided as part of this method, unique
         samples may require additional cleanup approaches to achieve the
         sensitivities stated in Table 1.
    3.3  It is necessary to remove diphenylamine frr- tr.e sample extract

-------
         prior to gas chromatography because it will interfere with the
         determination of N-nitrosodiphenylamine.  Removal is achieved if
         the sample is processed completely through one of the clean-up
         procedures detailed in the method.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1     Grab sample bottle - amber glass, 1-liter or 1-quart
                   volume.  French or Boston Round design is recommended.
                   The container must be washed and solvent rinsed before
                   use to minimize interferences.
         4.1.2     Bottle caps - Threaded to screw on to the sample
                   bottles.  Caps must be lined with Teflon.  Foil may be
                   substituted if sample is not corrosive.
         4.1.3     Compositing equipment - Automatic or manual compositing
                   system. Must incorporate glass sample containers for the
                   collection of a minimum of 250 ml.  Sample containers
                   must be kept refrigerated during sampling.  No tygon or
                   rubber tubing may be used in the system.
    4.2  Separatory funnels - 2000 ml and 250 ml, with Teflon stopcock.
    4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
         frit.
    4.4  Kuderna-Danish (K-D) Apparatus
         4.4.1     Concentrator tube - 10 ml, graduated (Kontes
                   K-570050-1025 or equivalent).  Calibration must be
                   checked.  Grc-^d glass stopper (size 19/22 joint) is used
                   to prevent ev;~?ration of extracts.

-------
         4.4.2  Evaporative flask  - 500 ml  (Kontes K-57001-0500 or equiva-
                lent).   Attach  to  concentrator tube with springs.  (Kontes
                K-662750-0012).
         4.4.3  Snyder  column -  three-ball  macro (Kontes K503000-0121 or
                equivalent).
         4.4.4-  Snyder  column -  two-ball micro (Kontes K-569001-0219 or
                equivalent).
         4.4.5  Boiling chips -  solvent extracted, approximately 10/40 mesh.
    4.5  Water bath - Heated, with concentric ring cover, capable of temper-
         ature control  ( + 2°C).  The bath should be used in a hood.
    4.6  Gas chromatograph - Analytical system complete with gas chromato-
         graph suitable for on-column injection and all required accessories
         including nitrogen-phosphorus or reductive Hall detector, column
         supplies, recorder, gases, syringes.  A data system for measuring
         peak areas is  recommended.
    4.7  Chromatographic column  -  Pyrex (approximately 300 mm long x 10 mm
         ID) with coarse fritted disc at bottom and Teflon stopcock (Kontes
         K-420540-0213  or equivalent).
    4.8  Chromatographic column  -  Pyrex (approximately 400 mm long x 22 mm
         ID) with coarse fritted disc at bottom and Teflon stopcock (Kontes
        \
         K-420540-0234  or equivalent).
5.  Reagents
    5.1  Preservatives:
         5.1.1  Sodium  hydroxide - (ACS) 10 N in distilled water.
         5.1.2  Sulfuric acid  - (ACS) Mix equal volumes of cone. H2S04
                with distilled  water.

                                '00

-------
     5.1.3  Sodium thiosulfate - (ACS) Granular.
5.2  Methylene chloride - Pesticide quality or equivalent.
5.3  Sodium Sulfate - (ACS) Granular, anhydrous (purified by heating at
     400°C for 4 hrs. in a shallow tray).
5.4  Stock standards - Prepare stock standard solutions at a concentra-
     tion of 1.00 ug/ul by dissolving 0.100 grams of assayed reference
     material in pesticide quality isooctane or other appropriate
     solvent and diluting to volume in a 100 ml ground glass stoppered
     volumetric flask.  The stock solution is transferred to ground
     glass stoppered reagent bottles, stored in a refrigerator, and
     checked frequently for signs of degradation or evaporation,
     especially just prior to preparing working standards from them.
5.5  Methyl alcohol, pentane, acetone - Pesticide quality or equivalent.
5.6  Diethyl Ether - Nanograde, redistilled in glass if necessary.
     5.6.1  Must be free of peroxides as indicated by EM Quant test
            strips.  (Test strips are available from EM Laboratories,
            Inc., 500 Executive Blvd., Elmsford, N.Y.  10523.)
     5.6.2  Procedures recommended for removal of peroxides are provided
            with the test strips.  After cleanup, 20 ml ethyl alcohol
            preservative must be added to each liter of ether.
5.7  Florisil - PR grade (60/100 mesh); purchase rjtivated at 1250°F
     and store in dark in glass containers witr r' — s stoppers or
     foil-lined screw caps.  Before use, activate ^;n batch at least 16
     hours at 130°C in a foil covered glass ccr'j'-~*r.
5.8  Alumina - Activity Super I, Basic, W20G ser.'er ^ICN Life Sciences
     Group, No. 404571).
                        TIT-

-------
    5.9  Hydrochloric acid, 10%-(ACS) Add one volume of cone. HC1 to nine
         volumes distilled water.
6.  Calibration
    6.1  Prepare calibration standards that contain the compounds of
         interest, either singly or mixed together.  The standards should be
       »
         prepared at concentrations covering two or more orders of magnitude
         that will completely bracket the working range of the chromato-
         graphic system.  If the sensitivity of the detection system can be
         calculated from Table I as 100 ug/1 in the final extract, for
         example, prepare standards at 10 ug/1, 50 ug/1, 100 ug/1, 500 ug/1,
         etc. so that injections of 1-5 ul of each calibration standard will
         define the linearity of the detector in the working range.
    6.2  Assemble the necessary gas chromatographic apparatus and establish
         operating parameters equivalent to those indicated  in Table I.  By
         injecting calibration standards, establish the sensitivity limit of
         the detector and the linear range of the analytical system for each
         compound.
    6.3  Before using any cleanup procedure, the analyst must process  a
         series of calibration standards through the procedure to validate
         elution patterns and the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing  any samples, the analyst should demonstrate
         through the analysis of a distilled water method blank, that  all
         glassware and  reagents  are  interference-free.  Each time a set of
         samples is extracted or there is a change in reagents, a method

-------
         blank should be processed as a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality assurance practices should be used with this
         method.   Field replicates should be collected to validate the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification of a peak on
         the chromatogram, confirmatory techniques such as mass spectroscopy
         should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prewashed with sample before collection.  Composite samples
         should be collected in refrigerated glass containers in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of collec-
         tion until extraction.  Chemical preservatives should not be used
         in the field unless more than 24 hours will elapse before delivery
         to the.laboratory.  If the samples will not be extracted within 48
         hours of collection, they must be preserved as follows:
         8.2.1  Add 35 mg of sodium thiosulfate per part per million of free
                chlorine per liter of sample.
         8.2.2  Adjust the pH of the water sample to pH 7 to 10 using sodium
                hydroxide or sulfuric acid.  Record the volume of acid or
                base added.

-------
    8.3  All samples must be extracted within 7 days and completely  analyzed
         within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for  later
         determination of sample volume.  Pour the entire sample into  a
         two-liter separatory funnel.  Check the pH of the sample with
         wide-range pH paper and adjust to within the range of 7 to  10 with
         sodium hydroxide or sulfuric acid.
    9.2  Add 60 ml methylene chloride to the sample bottle, seal, and  shake
         30 seconds to rinse the inner walls.  Transfer the solvent  into  the
         separatory funnel, and extract the sample by shaking the funnel  for
         two minutes with periodic venting to release vapor pressure.  Allow
         the organic layer to separate from the water phase for a minimum of
         ten minutes.  If the emulsion interface between layers is more than
         one-third the size of the solvent layer, the analyst must employ
         mechanical techniques to complete the phase separation.  The
         optimum technique depends upon the sample, but may include
         stirring, filtration of the  emulsion through glass wool, or
         centrifugation.  Collect the methylene chloride extract  in  a  250-ml
         separatory funnel.
    9.3  Add a  second 60-ml volume of methylene chloride to the sample
         bottle and complete the extraction procedure a second time,
         combining the extracts  in the 250-ml separatory funnel.
    9.4  Perform  a third  extraction  in the same manner.  Add  10 ml  of  10%
         HC1 solution to  the combined extracts  and  shake for  2 minutes.
         Allow  the  layers to separate.  Drain the methylene chloride layer

-------
     through a drying column containing 3-4 inches of anhydrous sodium
     sulfate, and collect it in a 500-ml Kuderna-Danish (K-D) flask
     equipped with a 10 ml concentrator tube.  Rinse the column with
     20-30 ml methylene chloride to complete the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus on a hot
     water bath (60-65°C) so that the concentrator tube is partially
     immersed in the hot water, and the entire lower rounded surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distillation
     the balls of the column will actively chatter but the chambers will
     not flood.  Because of the volatility of N-nitrosodimethyl amine,
     K-D concentration must be carefully carried out.  When the apparent
     volume of liquid reaches 1 ml, remove the K-D apparatus and allow
     it to drain for at least 10 minutes while cooling.  Remove the
     Snyder column and rinse the flask and its lower joint into the
     concentrator tube with 1-2 ml of methylene chloride.  A 5-ml
     syringe is recommended for this operation.  Unless the entire
     extract will be subjected to a cleanup operation (Section 10),
     adjust the extract volume to 10.0 ml with methylene chloride, add
     stopper, and refrigerate.
9.6  If the sample is being analyzed for N-nitrosodiphenylamine, the
     analyst must immediately proceed with one of the cleanup methods in

-------
         Section  10 to  remove potential  diphenylamine  interference.   Depend-
         ing upon the sensitivity requirement  for  the  analysis,  the  analyst
         may use  the entire  extract  for  this cleanup  as  described, or adjust
         the extract volume  to  10.0  ml with methylene  chloride  and pipet a 2
         ml aliquot onto  the column  in 10.2.2  or  10.3.3.
    9.7   If N-nitrosodiphenylamine is of no  interest,  the  analyst must
         choose between proceeding directly  to Section II,  or submitting the
         extract  to a cleanup procedure  before gas chromatography.   A
         solvent  exchange from  methylene chloride  to  methyl  alcohol  is
         required for direct gas chromatography.   Once the  entire extract is
         in methyl  alcohol  it cannot be  treated to either  of the cleanup
         procedures in  Section  10.   Therefore, in  the  absence of previous
         experience with  the sample  matrix,  the analyst  should  remove a 2.0
         ml aliquot of  the extract for gas chromatography  and retain the
         remainder  for  cleanup  if  required  later.
    9.8   Determine  the  original sample volume  by  refilling the  sample bottle
         to the mark  and  transferring  the liquid  to a 1000 ml graduated
         cylinder.  Record the  sample  volume to the nearest 5 ml.
10.  Cleanup and  Separation
    10.1  If the  entire  extract  is  to be  cleaned up by one  of the following
         procedures,  it must be concentrated to 2.0 ml.   To the concentrator
         tube  in  9.5,  add a clean  boiling chip and attach  a two-ball
         micro-Snyder  column.   Prewet  the column  by adding about 0.5 ml
         methylene  chloride to  the top.   Place the K-D apparatus on  a
         steaming hot   (60-65°C) water  bath so that the concentrator  tube
         is partially immersed  in  the  hot water.   Adjust the vertical

-------
     position of the apparatus and the water temperature as required to
     complete the concentration in 5-10 minutes.   At the proper rate of
     distillation the balls of the column will  actively chatter but the
     chambers will  not flood.   When the apparent  volume of liquid
     reaches about 0.5 ml,  remove the K-D and allow it to drain for at
                                                            1          *
     least  10 minutes while cooling.   Remove the  micro- Snyder column and
     rinse  its lower joint  into the concentrator  tube with 0.2 ml of
     methylene chloride.  Adjust the final volume to 2.0 ml and proceed
     with one of the following cleanup procedures.
10.2 Florisil Column Cleanup for Nitrosamines
     10.2.1  Place 22g of  activated Florisil in  a  22 mm' ID chromato-
            graphic column.  After settling the Florisil by tapping the
            column, add about a 5 mm layer of anhydrous granular sodium
            sulfate to the  top.
     10.2.2  Preelute the  column, after cooling, with 40 ml of 15% ethyl
            ether/85% pentane.  Discard the eluate and just prior to
            exposure of the sodium sulfate layer  to air, quantitatively
            transfer 2.0  ml of sample extract into the column by decan-
            tation using  an additional 2 ml of  pentane to complete the
            transfer.
     10.2.3  Perform the first elution with 90 ml  of 15% ethyl ether/85%
            pentane (V/V) and discard the eluate.  This fraction will
            contain any diphenylamine.
     10.2.4  Perform the second elution with 100 ml of 5% acetone/95%
            ethyl ether (V/V)  and collect the eluate in a 500-ml K-D
            flask equipped  with a 10-ml concentrator tube.  This
                          ur

-------
            fraction will contain all of the nitrosamines.
     10.2.5 Add 15 ml of methanol to the collected eluate and concen-
            trate as in 9.5 at 70-75°C, substituting pentane for
            methylene chloride.
     10.2.6 Analyze by gas chromatography.
10.3 Alumina Column Cleanup for Nitrosamines
     10.3.1 Place lOOg of alumina, as it comes from the manufacturer,
            into a 500 ml reagent bottle and add 2 ml of distilled
            water, which is free of nitrosamines and interferences.  Mix
            the alumina preparation thoroughly by shaking or rolling for
            10 minutes and let it stand for at least 2 hours.  The
            preparation should be homogeneous before use.  Keep the
            bottle sealed tightly to ensure proper activity.
     10.3.2 Place 12 grams of the alumina preparation into a 10 mm ID
            chromatographic column and tap the column to settle the
            alumina.  Add 1-2 cm of anhydrous sodium sulfate to the top
            of the alumina.
     10.3.3 Preelute the column with 10 ml of 30% ethyl ether/70%
            pentane (V/V).  Discard the eluate (about 2 ml) and, just
            prior to exposure of the sodium sulfate layer the air,
            transfer 2.0 ml of sample extract onto the column by decan-
            tation using an additional 2 ml of pentane to complete the
            transfer.
     10.3.4 Just prior to exposure of the sodium sulfate layer to the
            air, add 70 ml of 2?« ethyl ether/70% pentane.  Discard the
            first 10 ml of elus."- - Jt collect the rest of the eluate in
                            S

-------
                a 500-ml  K-D flask  equipped with a 10 ml  concentrator tube.
                This  fraction contains N-nitrosodiphenylamine and probably a
                small  amount of N-nitrosodi-n-propylamine.
         10.-3.5 Next  elute the column with 60 ml of 50* ethyl ether/50%
                pentane,  collecting the eluate in a second  K-D flask
                equipped  with a 10  ml concentrator tube.   Add 15 ml methyl
                alcohol  to the K-D.  This fraction will contain
                N-nitrosodimethylamine, most of the
                N-nitrosodi-n-propylamine and any diphenylamine.
         10.3.6 Concentrate both fractions as in 9.5 substituting pentane
                for methylene chloride.
         10.3.7 Analyze by gas chromatography.
11.  Gas Chromatography
    11.1  N-nitrosodiphenylamine completely reacts to form diphenylamine at
          normal operating temperatures of the GC injection port.  There-
          fore, N-nitrosodiphenylamine is actually chromatographed and
          detected as diphenylamine.  The determination of either of the
          compounds in the original sample would be uncertain without the
          use of one  of the previous cleanup procedures which separate the
          two compounds.
    11.2  Table I summarizes some recommended gas chromatographic column
          materials and operating conditions for the instrument.  Included
          in this table are estimated retention times and sensitivities that
          should be achieved by this method.  Examples of the separations
          achieved by the primary column are shown in Figures 1 and 2.
          Calibrate the system daily with a minimum of three injections of

-------
      calibration standards.
11.3  If the extract has not been submitted to one of the cleanup proce-
      dures in Section 10, it is necessary to exchange the solvent from
      methylene chloride to methyl alcohol before the thermionic detec-
      tor can be used.  To a 1-10 ml volume of methylene chloride
      extract in a concentrrator tube, add 2 ml methyl alcohol, and a
      clean boiling chip.  Attach a two-ball micro-Snyder column.
      Prewet the column by adding about 0.5 ml methylene chloride
      through the top.  Place the K-D apparatus on a boiling water bath
      so that the concentrator tube is partially immersed in the hot
      water.  Adjust the vertical position and insulate the apparatus as
      necessary to complete the concentration in 5-10 minutes.  At the
      proper rate of distillation the balls of the column will actively
      chatter but the chambers will not flood.  When the apparent volume
      of liquid reaches about 0.5 ml, remove the K-D and allow it to
      drain for at least 10 minutes while cooling.  Remove the
      micro-Snyder column and rinse its lower joint into the concen-
      trator tube with 0.2 ml of methyl alcohol.  Adjust the final
      volume to 2.0 ml.
11.4  Inject 2-5 ul of the  sample extract using the solvent-flush
      technique.  Smaller (1.0 ul) volumes can be injected if automatic
      devices are employed.  Record the volume injected to the nearest
      0.05 ul, and the resulting peak size, in area units.
11.5  If the peak area exceeds the  linear range of the  system, dilute
      the extract and reanalyze.
11.6  If the peak area measurement  is prevented by the  presence  of

-------
          interferences, further cleanup is required.
12.  Calculations
    12.1  Determine the concentration of individual compounds according to
          the formula:
         Concentration, ug/1  -'(A)  (B) (Vt)
                                 (Vi)  (Vs)
         where A = Calibration factor for chromatographic system, in
                   nanograms  material per area unit.
               B = Peak size  in injection of sample extract, in area units
               V. = volume of extract injected (ul)
               V. = Volume of total extract (ul)
               V  = Volume of water extracted (ml)
    12.2  Report results in micrograms per liter  without correction for
          recovery data.  When duplicate and spiked samples are analyzed,
          all data obtained should  be reported.
13.  Accuracy and Precision
    The U.S. EPA Environmental Monitoring and Support Laboratory in
    Cincinnati is in the process of conducting an interlaboratroy method
    study to determine the accuracy and precision of this test procedure.

                                 BIBLIOGRAPHY
"Development and Application  of Test Procedures for Specific Organic Toxic
Substances in Wastewaters.  Category 5 - Nitrosamines," Report for EPA .
Contract 68-03-2606 (In preparation).
                                     iu

-------
                                   TABLE I

                      GAS CHROMATOGRAPHY OF NITROSAMINES
                          Retention Time (min.)
                                                               Detection
Nitrosamine
N-nitrosodidimethylamine
N-nitrosodi-n-dipropylamine
N-nitrosodidiphenylamine
Column 1
4.1
12.1
12.8*
Column 2
0.88
4.2
6.4**
Limit (ug/1)
0.3
0.5
1.0
Column 1 conditions:  Chromosorb WAW 80/100 mesh coated with 10% Carbowax
    20M/2% KOH packed in a 180 on long x 4 mm ID glass column with helium
    carrier gas at 40 ml/min flow rate.   Isothermal  column temperature is
    110°C except where * indicates 220°C.

Column 2 conditions:  Supelcoport 100/120 mesh coated with 10% SP-2250
    packed in a 180 cm long x 4 mm ID glass column with helium carrier gas
    at 40 ml/mi^flow rate.  Isothermal  column temperature is 120°C except
    where ** indicates 210°C.

Detection limit is calculated from the minimum detectable GC response being
    equal to five times the GC background noise, assuming a 10 ml final
    volume of the 1 liter sample extract, and assuming a GC injection of 5
    microliters.  A nitrogen-phosphorus detector was used to collect this
    data, but a Thermal Energy Analyzer exhibited equivalent sensitivity.
                            Ml

-------
COLUMN:  10% CARBOWAX 20M + 2% KOH OM CHROMOSORB W-AW

TEMPERATURE: 110°

DETECTOR: PHOSPHORUS/NITROGEN
                   Q_
                   o
                   cc
                   a.
                   O
                   CO

                   O
                   cc
    24  6   8  10  12  14


  RETENTION TIME-MINUTES


 Figure 1. Gas chromatogram of nitrosamines

-------
 COLUMN:  10% CARBOWAX 20M •»- 2% KOH ON CHROMOSORB YV-AW
 TEMPERATURE: 220°C.
 DETECTOR: PHOSPHORUS/NITROGEN
 0  2   4   6  8  10  12  14  16  18

         RETENTION TIME-MINUTES

Figure  2.  Gas chromatogram of N-nitrosodiphenylamine
          as diphenylamine

-------
                     ORGANOCHLORINE PESTICIDES AND PCB's
                                  METHOD 608
1.   Scope and Application
    1.1   This method covers the determination of certain organochlorine
         pesticides and polychlorinated biphenyls (PCBs).   The following
         parameters may be determined by this method:
         Parameter
         Aldrin
         a-BHC
         b-BHC
         d-BHC
         g-BHC
        v
         Chlordane
         4,4'-DDD
         4,4'-DDE
         4,4'-DDT
         Dieldrin
         Endosulfan  I
         Endosulfan  II
         Endosulfan  Sulfate
STORET No.
  39330
  39337
  39338
  39259
  39340
  39350
  39310
  39320
  39300
  39380
  34361
  34356
  34351
Parameter
Endrin
Endrin Aldehyde
Heptachlor
Heptachlor Epoxide
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
STORET No
   39390
   34366
   39410
   39420
   39400
   34671
   39488
   39492
   39496
   39500
   39504
   39508
    1.2   This  method  is  applicable to  the determination of these compounds
         in  municipal  and  industrial discharges.   It  is designed to be used
         to  meet  the  monitoring  requirements  of  the National  Pollutant
         Discharge  Elimination System  (NPDES).   As such,  it presupposes a
                                  -  113

-------
         high expectation of finding the specific compounds of interest.  If
         the user is attempting to screen samples for any or all  of the
         compounds above, he must develop independent protocols for the
         verification of identity.
    1.3  The sensitivity of this method is usually dependent upon the level
         of interferences rather than instrumental limitations.  The limits
         of detection listed in Table I represent sensitivities that can be
         achieved in wastewaters in the absence of interferences..
    1.4  This method is recommended for use only by experienced residue
         analysts or under the close supervision of such qualified persons.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
         using separatory funnel techniques.   The extract is dried and
         concentrated to a volume of 10 ml or less.  Chromatographic
         conditions are described which allow for the accurate measurement
         of the compounds in the extract.
    2.2  If interferences are encountered, the method provides selected
         general purpose cleanup procedures to aid the analyst in their
         elimination.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
                             UT-

-------
     in all-glass systems may be required.
3.2  Interferences coextracted from the samples will vary considerably
     from source to source, depending upon the diversity of the
     industrial complex or municipality being sampled.  While general
     cleanup techniques are provided as part of this method, unique
     samples may require additional cleanup approaches to achieve the
     sensitivities stated in Table 1.
3.3  Glassware must be scrupulously clean.  Clean all glassware as soon
     as possible after use by rinsing with the last solvent used.  This
     should be followed by detergent washing in hot water.  Rinse with
     tap water, distilled water, acetone and finally pesticide quality
     hexane.  Heavily contaminated glassware may require treatment in a
     muffle furnace at 400°C for 15 to 30 minutes.  Some high boiling
     materials, such as PCBs, may not be eliminated by this treatment.
     Volumetric ware should not be heated in a muffle furnace.
     Glassware should be sealed/stored in a clean environment
     immediately after drying or cooling to prevent any accumulation of
     dust or other contaminants.  Store inverted or capped with aluminum
     foil.
3.4  Interferences by phthalate esters can pose a major problem in
    .pesticide analysis.  These materials elute in the 15% and 50%
     fractions of the Florisil cleanup.  They usually can be minimized
     by avoiding contact with any plastic materials.  The contamination
     from phthalate esters can be completely eliminated with the use of
     a microcoulometric or electrolytic conductivet detector.
                        77T - 11

-------
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1     Grab sample bottle - amber glass, 1-liter or 1-quart
                   volume.  French or Boston Round design is recommended.
                   The container must be washed and solvent rinsed before
                   use to minimize interferences.
         4.1.2     Bottle caps - Threaded to screw on to the sample
                   bottles.  Caps must be lined with Teflon.  Foil may be
                   substituted if sample is not corrosive.
         4.1.3     Compositing equipment - Automatic or manual compositing
                   system. Must incorporate glass sample containers for the
                   collection of a minimum of 250 ml.  Sample containers
                   must be kept refrigerated during sampling.  No tygon or
                   rubber tubing may be used in the system.
    4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
    4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
         frit.
    4.4  Kuderna-Danish (K-D) Apparatus
         4.4.1     Concentrator tube - 10 ml, graduated (Kontes
                   K-570050-1025 or equivalent).  Calibration must be
                   checked at 1.0 and 10.0 ml level.  Ground glass stopper
                   (size  19/22 joint) is used to prevent  evaporation of
                   extracts.
         4.4.2     Evaporative flask - 500 ml (Kontes K-57001-0500 or
                   equivalent).  Attach to concentrator tube with springs.
                   (Kontes K-662750-0012).
                           J7T

-------
         4.4.3     Snyder column - three-ball  macro (Kontes K503000-0121 or
                   equivalent).
         4.4.4     Boiling chips - extracted,  approximately 10/40 mesh.
    4.5  Water bath - Heated,  with concentric  ring cover,  capable of
         temperature control  (+ 2°C).   The bath should be  used in a hood.
    4.6  Gas chromatograph -  Analytical  system complete with gas
         chromatograph suitable for on-column  injection and all requ-ired
         accessories including electron  capture or halogen-specific
         detector,  column supplies, recorder,  gases, syringes.  A data
         system for measuring peak areas is recommended.
    4.7  Chromatographic column - Pyrex, 400 mm x 25 mm OD, with coarse
         fritted plate and Teflon stopcock (Kontes K-42054-213 or
         equivalent).
5.  Reagents
    5.1  Preservatives:
         5.1.1     Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2     Sulfuric acid (1+1) - (ACS) Mix equal volumes of cone.
                   H2S04 with distilled water.
    5.2  Methylene chloride - Pesticide quality or equivalent.
    5.3  Sodium Sulfate - (ACS) Granular, anhydrous (purified by heating at
         40t)°C for 4 hrs. in a shallow tray).
    5.4  Stock standards - Prepare stock standard solutions at a
         concentration of 1.00 ug/ul by dissolving 0.100 grams of assayed
         reference material in pesticide quality isooctane or other
         appropriate solvent and diluting to volume in a 100 ml ground glass
         stoppered volumetric flask.  The stock solution is transferred to
                                  "7

-------
         ground  glass stoppered reagent bottles,  stored in a refrigerator,
         and checked frequently for signs of degradation or evaporation,
         especially just prior to preparing working standards from them.
    5.5   Boiling chips - Hengar granules (Hengar  Co.;  Fisher Co.)  or
         equivalent.
    5.6   Mercury -  triple distilled.
    5.7   Aluminum oxide - basic or neutral, active.
    5.8   Hexane  - pesticide residue analysis grade.
    5.9   Isooctane  (2,2,4-trimethyl pentane) - pesticide residue analysis
         grade.
    5.10 Acetone -  pesticide residue analysis grade.
    5.11 Diethyl ether - Nanograde, redistilled in glass if necessary.
         5.11.1     Must be free of peroxides as indicated by EM Quant test
                   strips (Test strips are available from EM Laboratories,
                   Inc., 500 Executive Blvd., Elmsford, N.Y., 10523).
         5.11.2     Procedures recommended for removal  of peroxides are
                   provided with the test strips.  After cleanup 20 ml  ethyl
                   alcohol preservative must be added to each liter of ether.
    5.12 Florisil - PR grade (60/100 mesh); purchase activated at 1250°F
         and store  in glass containers with glass stoppers or foil-lined
         screw caps.  Before use activate each batch at least 16 hours at
         130°C in a foil covered glass container.
6.  Calibration
    6.1   Prepare calibration standards that contain the compounds of
         interest,  either singly or mixed together.  The standards should be
         prepared at concentrations covering two or more orders of magnitude
                            ID

-------
         that will completely bracket the working range of the
         chromatographic system.  If the sensitivity of the detection system
         can be calculated from Table I as 100 ug/1 in the final extract,
         for example, prepare standards at 10 ug/1, 50 ug/1, 100 ug/1, 500
         ug/1, etc., so that injections of 1-5 ul of each calibration
         standard will define the linearity of the detector in the working
         range.
    6.2  Assemble the necessary gas chromatographic apparatus and establish
         operating parameters equivalent to those indicated in Table I.  By
         injecting calibration standards, establish the sensitivity limit of
         the detector and the linear range of the analytical system for each
         compound.
    6.3  The cleanup procedure in Section 10 utilizes Florisil
         chromatography.  Florisil from different batches rt«* sources may
         vary in absorption capacity.  To standardize the amount of Florisil
         which is used, the use of lauric acid value (Mills, 1968) is
         suggested.  The referenced procedure determines the adsorption from
         hexane solution of lauric acid (mg)  per gram Florisil.  The amount
         of Florisil to be used for each column is calculated by dividing
         this factor into 110 and multiplying by 20 grams.
    6.4  Before using any cleanup procedure,  the analyst must process a
         series of calibration standards through the procedure to validate
         elution patterns and the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any si."les,  the analyst should demonstrate
         through the analysis of i distilled  water method blank, that all

                               HI    '«?

-------
         glassware and reagents are interference-free.  Each time a set of
         samples is extracted or there is a change in reagents, a method
         blank should be processed as a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality assurance practices should be used with this
         method.  Field replicates should be collected to validate the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification of a peak on
         the chromatogram, confirmatory techniques such as mass spectroscopy
         should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prewashed with sample before collection.  Composite samples
         should be collected in refrigerated glass containers  in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of  contamination.
    8.2  The samples must be iced or refrigerated from the time of
         collection until extraction.  Chemical preservatives  should not be
         used in the field unless more than 24 hours will elapse before
         delivery to the laboratory.  If the samples will not  be extracted
         within 48 hours of collection, the sample should be adjusted to a
         pH range of 6.0-8.0 with sodium hydroxide or sulfuric acid.
    8.3  All samples must be extracted within 7 days  and completely analyzed

-------
         within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for later
         determination of sample volume.  Pour the entire sample into a
         two-liter separatory funnel.  Check the pH of the sample with
         wide-range pH paper and adjust to within the range of 5-9 with
         sodium hydroxide or sulfuHc acid.
    9.2  Add 60 ml methylene chloride to the sample bottle, seal, and shake
         30 seconds to rinse the inner walls.  Transfer the solvent into the
         separatory funnel, and extract the sample by shaking the funnel for
         two minutes with periodic venting to release vapor pressure.  Allow
         the organic layer to separate from the water phase for a minimum of
         ten minutes.  If the emulsion interface between layers is more than
         one-third the size of the solvent layer, the analyst must employ
         mechanical techniques to complete the phase separation.  The
         optimum technique depends upon the sample, but may include
         stirring, filtration of the emulsion through glass wool, or
         centrifugation.  Collect the methylene chloride extract in a 250-ml
         Erlenmeyer flask.
    9.3  Add a second 60-ml volume of methylene chloride to the sample
 «       bottle and complete the extraction procedure a second time,
         combining the extracts in the Erlenmeyer flask.
    9.4  Perform a third extraction in the same manner.  Pour the combined
         extract through a drying column containing 3-4 inches of anhydrous
         sodium sulfate, and collect it in a 500-ml Kuderna-Danish (K-D)
         flask equipped with a 10 ml concentrator tube.  Rinse the

-------
     Erlenmeyer flask and column with 20-30 ml methylene chloride to
     complete the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus on a hot
     water bath (60-65°C) so that the concentrator tube is partially
     immersed in the hot water, and the entire lower rounded surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distillation
     the balls of the column will actively chatter but the chambers will
     not flood.  When the apparent volume of liquid reaches 1 ml, remove
     the K-D apparatus and allow it to drain for at least 10 minutes
     while cooling.
9.6  Increase the temperature of the hot water bath to about 80°C.
     Momentarily remove the Snyder column, add 50 ml of hexane and a new
     boiling chip and reattach the Snyder column.  Pour about 1 ml of
     hexane into the top of the Snyder column and concentrate the sol-
     vent extract as before.  The elapsed time of concentration should
     be 5 to 10 minutes.  When the apparent volume of liquid reaches 1
     ml, remove the K-D apparatus and allow it to drain at least 10
     minutes while cooling.  Remove the Snyder column and rinse the
     flask and its lower joint into the concentrator tube with 1-2 ml of
     hexane, and adjust the volume to 10 ml.  A 5-ml syringe is
     recommended for this operation.  Stopper the concentrator tube and
     store refrigerated if further processing will not be performed

-------
         Immediately.  If the sample extract requires no further cleanup,
         proceed with gas chromatographic analysis.  If the sample requires
         cleanup, proceed to Section 10.
    9.7  Determine the original  sample volume by refilling the sample bottle
         to the mark and transferring the liquid to a 1000 ml graduated
         cylinder.  Record the sample volume to the nearest 5 ml.
10.  Cleanup and Separation
    10.1 Cleanup procedures are  used to extend the sensitivity of a method
         by minimizing or eliminating interferences that mask or otherwise
         disfigure the gas chromatographic response to the pesticides and
         PCB's.  The Florisil column allows for a select fractionation of
         the compounds and will  eliminate polar materials.  Elemental sulfur
         interferes with the electron capture gas chromatography of certain
         pesticides but can be removed by the techniques described below.
    10.2 Florisil Column Cleanup
         10.2.1 Add a weight of  Florisil, (nominally 21g,) predetermined by
                calibration (6.3,  6.4), to a chromatographic column.  Settle
                the Florisil by  tapping the column.  Add sodium sulfate to
                the top of the Florisil to form a layer 1-2 cm deep.  Add 60
                ml of hexane to  wet and rinse the sodium sulfate and
                Florisil.  Just  prior to exposure of the sodium sulfate to
                air, stop the elution of the hexane by closing the stopcock
                on the chromatography column.  Discard the eluate.
         10.2.2 Adjust  the sample extract volume to 10 ml and transfer it
                from the K-D concentrator tube to the Florisil

-------
               column.  Rinse the tube twice with  1-2 ml  hexane,  adding
               each rinse to the column.
        10.2.3 Place a 500 ml K-D flask  and clean  concentrator  tube  under
               the chromatography column.  Drain the column  into  the flask
               until the sodium sulfate  layer  is nearly  exposed.   Elute the
               column with 200 ml of 6%  ethyl  ether  in hexane  (Fraction 1)
               using a drip rate of about  5 ml/min.  Remove  the K-D  flask
               and set aside for later concentration.  Elute the  column
               again, using 200 ml of  15%  ethyl ether  in hexane (Fraction
               2), into a second K-D flask.  Perform the third  elution
               using 200 ml of 50% ethyl ether in  hexane (Fraction 3).   The
               elution patterns.for the  pesticides and PCB's are  shown  in
               Table II.
        10.2.4 Concentrate the eluates by  standard K-D techniques (9.5),
               substituting hexane for the glassware rinses  and using the
               water bath at  about 85°C.   Adjust final volume  to  10  ml
               with hexane.   Analyze by  gas chromatography.
    10.3 Elemental  sulfur will  usually  elute entirely in  Fraction  1.   To
        remove sulfur  interference from  this fraction  or the original
        extract,  pi pet  1.00 ml of  the  concentrated extract into a clean
        concentrator tube or  Teflon-sealed vial.   Add  1-3 drops of mercury
        and  seal.  Agitate the contents  of the vial  for  15-30  seconds.
        Place the vial  in  an  upright position  on  a reciprocal  laboratory
        shaker  and shake  for  2 hours.   Analyze by  gas  chrx -;tography.
11.  Gas Chromatography
    11.1 Table I  summarizes  some  recommended  gas chromatoc'  - ic column

-------
         materials  and operating conditions for the instrument.  Included in
         this  table are estimated retention times and sensitivities that
         should be  achieved by this method.  Examples of the separations
         achieved by these columns are shown in Figures 1 through 10.
         Calibrate  the system daily with a minimum of three injections of
         calibration standards.
    11.2 Inject 2-5 ul of the sample extract using the solvent-flush
         technique.  Smaller (1.0 ul)  volumes can be injected if automatic
         devices are employed.  Record the volume injected to the nearest
         0.05  ul, and the resulting peak size, in area units.
    11.3 If the peak area exceeds the linear range of the system, dilute the
         extract and reanalyze.
    11.4 If the peak area measurement is prevented by the presence of
         interferences, further cleanup is required.
12.  Calculations
    12.1 Determine  the concentration of individual compounds according to
         the formula:
         Concentration, ug/1 = (A) (B) (Vt)
                               (Vi)  (Vs)
         where A =  Calibration factor for chromatographic system, in
                   nanograms material  per area unit.
               B =  Peak size in injection of sample extract, in area units
               V^ = volume of extract injected (ul)
               Vt = Volume of total extract (ul)
               V  = Volume of water extracted (ml)
    12.2 Report results in micrograms per liter without correction for

-------
         recovery data.  When duplicate and spiked samples are analyzed, all
         data obtained should be reported.
13.  Accuracy and Precision
    13.1 The U.S. EPA Environmental Monitoring and Support Laboratory in
         Cincinnati is in the process of conducting an interlaboratory
         method study to determine the accuracy and precision of this test
         procedure.

                                 BIBLIOGRAPHY
1.  "Development and Application of Test Procedures for Specific Organic
    Toxic Substances in Wastewaters.  Category 10-Pesticides and PCB's."
    Report for EPA Contract 68-03-2606.
    Mills, P. A., "Variation of Florisil Activity:  Simple Method for
    Measuring Absorbent Capacity and Its Use in Standardizing Florisil
    Columns," Journal of the Association of Official Analytical Chemists,
    51, 29 (1968).

-------
                                   TABLE I

                  GAS CHROMATOGRAPHY OF PESTICIDES AND PCB's
Parameter
Retention Time (min)
Column 1    Column 2
Detection Limit
     (ug/1)1
Aldrin
a-BHC
b-BHC
d-BHC
g-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
2.40
1.35
1.90
2.15
1.70
*
7.83
5.13
9.40
5.45
4.50
8.00
14.22
6.55
11.82
2.00
3.50
*
*
*
*
*
*
*
*
4.10
1.82
1.97
2.20
2.13
*
9.08
7.15
11.75
7.23
6.20
8.28
10.70
8.10
9.30
3.35
5.00
*
*
*
*
*
*
it
it
0.003
0.002
0.004
0.004
0.002
0.04
0.012
0.006
0.016
0.006
0.005
0.01
0.03
0.009
0.023
0.002
0.004
0.40
0.04
0.10
0.10
0.05
0.08
0.08
0.15
*   Multiple peak response.  See Figures 2-10.

**  Detection limit is calculated from the minimum detectable GC response
    being equal  to five times the GC background noise, assuming a 10 ml
    final volume of the 1 liter sample extract, and assuming a GC injection
    of 5 micro!iters.

Column 1 conditions:  Supelcoport 100/120 mesh coated with 1.5%
    SP-2250/1.95% SP-2401 packed in a 180 cm long x 4 mm ID glass column
    with 5% Methane/95% Argon carrier gas at 60 ml/min flow rate.  Column
    temperature is 200°C.

Column 2 conditions:  Supelcoport 100/120 mesh coated with 3% OV-1 in a 180
    cm long x 4 mm ID glass column with 5% Methane/95% Argon carrier gas at
    60 ml/min flow rate.  Column temperature is 200°C.

-------
                                   TABLE II

             DISTRIBUTION AND RECOVERY OF CHLORINATED PESTICIDES
                AND PCBs USING FLORISIL COLUMN CHROMATOGRAPHY
Parameter
Recovery (%) by Fraction*
1(6%)    2(15%)    3(50*)
Aldrin
a-BHC
b-BHC
d-BHC
g-BHC
Chlordane
4,4'-ODD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
100
100
 97
 98
100
100
 99
 98
100
  0
 37
  0
  0
  4
  0
100
100
 96
 97
 97
 95
 97
103
 90
 95
100
 64
  7
  0
 96
 68
 91
106

 26
*From:  "Development and Application of Test Procedures for Specific Organic
         Toxic Substances in Wastewaters.  Category 10-Pesticides and PCB's.
         Report for EPA Contract 68-03-2606."
                               .	.    ,  , ^~<
                              jli -  D:V

-------
   COLUMN:  1.5% SP-2250*
            1.95% SP-2401 DIM SUPELCOPORT
   TEMPERATURE:  200°C.
   DETECTOR: ELECTRON CAPTURE
4        8       12
 RETENTION TIME-MINUTES
                                  16
Figure 1.  Gas chromatogram of pesticides

-------
   COLUMN: 1.5% SP-2250*
            1.95% SP-2401 ON SUPELCOPORT
   TEMPERATURE: 200°C.
   DETECTOR: ELECTRON CAPTURE
        4        8       12
        RETENTION TIME-MINUTES
16
Figure 2.  Gas chromatogram of chlordane
     JL

-------
                      COLUMN: 1.5% SP-2250+
                               1.95?'« SP-2401 ON SUPELCOPORT
                      TEMPERATURE: 200'C.
                      DETECTOR: ELECTRON CAPTURE
   2       6      10      14      18      22
                 RETENTION TIME-MINUTES

Figure 3.  Gas chromatogram of toxaphene
26

-------
COLUMN:  1.5% SP-2250+ 1.95%
 EMPERATURE:  160aC.
DETECTOR: ELECTRON  CAPTURE
                          o SP-2401 ON SUPELCOPORT
    2       6       10      14       18
            RETENTION TIME-MINUTES

Figure 4.  Gas chromatogram of PCB-1016
                                         22

-------
  COLUMN: 1.5% SP-2250+ 1.95% SP-2401 DIM SUPELCOPORT
  TEMPERATURE: 160'C.
  DETECTOR: ELECTRON CAPTURE
   2       6       10       14       18
             RETENTION TIME-MINUTES

Figure 5.  Gas chromatogram of PCB-1221
22
                            - e

-------
    COLUMN:  1.5JS SP-2250* 1.95°i SP-2401 ON SUPELCOPORT
    TEMPERATURE:  160°C.
    DETECTOR:  ELECTRON  CAPTURE
    2        6        10       14       18
                 RETENTION TIME-MINUTES

Figure 6.  Gas chromatogram of PCB-1232
22
24
                  tT -

-------
    COLUMN:  1.5% SP-2250+1.95% SP-2401 ON SUPELCOPORT
    TEMPERATURE:  160°C.
    DETECTOR:  ELECTRON  CAPTURE
    2       6        10       14      18
               RETENTION TIME-MINUTES

Ffgure 7.  Gas chromatogram of PCB-1242
22

-------
    COLUMN: 1.5% SP-2250* 1.95% SP-2401 ON SUPELCOPORT
    TEMPERATURE:  160°C.
    DETECTOR:  ELECTRON CAPTURE
   2       6       10      14       18
                 RETENTION TIME-MINUTES

Figure 8.  Gas chromatogram of PCB-1248
22
26

-------
COLUMN:  1.5% SP-2250 + 1.95% SP-2401 ON SUPELCOPORT
TEMPERATURE:  200°C.
DETECTOR: ELECTRON CAPTURE
    2        6        10       14       18

               RETENTION TIME-MINUTES

Figure 9.  Gas chromatogram of PCB-1254
                                            22

-------
    COLUMN:  1.5% SP-2250+ 1.95% SP-2401 ON SUPELCOPORT
    TEMPERATURE:  200aC.
    DETECTOR: ELECTRON CAPTURE
    2       6       10      14      18      22
                   RETENTION TIME-MINUTES

Figure 10. Gas chromatogram of PCB-1260
26

-------
                        NITROAROMATICS AND ISOPHORONE
                                  METHOD 609
1.   Scope and Application
    1.1  This method covers the determination of certain nitroaromatics and
         isophorone.  The following parameters may be determined by this
         method:
         Parameter                                           STORET No.
         Isophorone                                           34408
         Nitrobenzene                                         34447
         2,4-Dinitrotoluene                                   34611
         2,6-Dinitrotoluene                                   34626
    1.2  This method is applicable to the determination of these compounds
         in municipal and industrial discharges.  It is designed to be used
         to meet-the monitoring requirements of the National Pollutant
         Discharge Elimination System (NPDES).  As such, it presupposes a
         high expectation of finding the specific compounds of interest.  If
         the user is attempting to screen samples for any or all of the
         compounds above, he must develop independent protocols for the
         verification of identity.
    1.3  The sensitivity of this method is usually dependent upon the level
         of interferences rather than instrumental limitations.  The limits
         of detection listed in Table I represent sensitivities that can be
         achieved in wastewaters in the absence of interferences.

-------
    1.4  This method is recommended for use only by experienced residue
         analysts or under the close supervision of such qualified persons.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
         using separatory funnel techniques.  The extract is dried and
         exchanged to toluene while being concentrated to 1.0 ml.
         Isophorone and nitrobenzene are measured by flame ionization gas
         chromatography.  The dinitrotoluenes are measured by electron
         capture GC.
    2.2  If interferences are encountered, the method provides a general
         purpose cleanup procedure to aid the analyst in their elimination.
3.  Interferences
    3.1  Solvents,, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
         in all-glass systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the
         industrial complex or municipality being sampled.  While general
         clean-up  techniques are provided as part of this method, unique
         samples may require additional cleanup approaches to achieve the
         sensitivities stated in Table I.
4.  Apparatus and  Materials

-------
4.1  Sampling equipment, for discrete or composite sampling.
     4.1.1     Grab sample bottle - amber glass, 1-liter or 1-quart
               volume.  French or Boston Round design is recommended.
               The container must be washed and solvent rinsed before
               use to minimize interferences.
     4.1.,2     Bottle caps - Threaded to screw on to the sample
               bottles.  Caps must be lined with Teflon.  Foil may be
               substituted if sample is not corrosive.
     4.1.3     Compositing equipment - Automatic or manual compositing
               system. Must incorporate glass sample containers for the
               collection of a minimum of 250 ml.  Sample containers
               must be kept refrigerated during sampling.  No tygon or
               rubber tubing may be used in the system.
4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
     frit.
4.4  Kuderna-Danish (K-D) Apparatus
     4.4.1     Concentrator tube - 10 ml, graduated (Kontes
               K-570050-1025 or equivalent).  Calibration must be
               checked.  Ground glass stopper (size 19/22 joint) is used
               to prevent evaporation of extracts.
     4.4.2     Evaporative flask - 500 ml (Kontes K-57001-0500 or
               equivalent).  Attach to concentrator tube with springs.
               (Kontes K-662750-0012).
     4.4.3     Snyder column - three-ball macro (Kontes K503000-0121 or
               equivalent).
                           O

-------
         4.4.4     Snyder column - two-ball  micro (Kontes K-569001-0219 or
                   equivalent).
         4.4.5     Boiling chips - solvent extracted, approximately 10/40
                   mesh.
    4.5"  Water bath - Heated,  with concentric ring cover, capable of
         temperature control  (+ 2°C).   The bath should be used 1n a hood.
    4.6  Gas chromatograph -  Analytical system complete with gas
         chromatograph suitable for on-column Injection and all required
         accessories including both electron capture and flame ionization
         detectors, column supplies, recorder, gases, syringes.  A data
         system for measuring peak areas is recommended.
    4.7  Chromatography column - 400 mm long x 10 mm ID, with coarse fritted
         plate on bottom and  Teflon stopcock.
5.  Reagents
    5.1  Preservatives:
         5.1.1     Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2     Sulfuric acid (1+1) - (ACS) Mix equal volumes of cone.
                   H2S04 with distilled water.
    5.2  Methylene chloride - Pesticide quality or equivalent.
    5.3  Sodium sulfate - (ACS) Granular, anhydrous (purified by heating at
         400°C for 4 hrs. in  a shallow tray).
    5.4  Stock standards - Prepare stock standard solutions at a
         concentration of 1.00 ug/ul by dissolving 0.100 grams of assayed
         reference material in pesticide quality isooctane or other
         appropriate solvent  and' diluting to volume in a 100 ml ground glass
         stoppered volumetric flask.  The stock solution is transferred to

-------
         ground glass stoppered reagent bottles, stored  in  a  refrigerator,
         and checked frequently for signs of degradation or evaporation,
         especially just prior to preparing working standards from them.
    5.5  Acetone, Hexane, Methanol, Toluene - pesticide quality or
         equivalent.
    5.6  Florisil - PR grade (60/100 mesh); purchase activated at 1250°F
         and store in glass containers with glass stoppers  or foil-lined
         screw caps.  Before use, activate each batch overnight at 200°C
         in glass containers loosely covered with foil.
6.  Calibration
    6.1  Prepare calibration standards that contain the compounds of
         interest, either singly or mixed together.  The standards should be
         prepared at concentrations covering two or more orders of magnitude
         that will completely bracket the working range of  the
         chromatographtc system,  if the sensitivity of the detection system
         can be calculated from Table I as 100 ug/1 in the  final extract,
         for example, prepare standards at 10 ug/1, 50 ug/1, 100 ug/1, 500
         ug/1, etc. so that injections of 1-5 ul of each calibration
         standard will define the linearity of the detector in the working
         range.
    6.2  Assemble the necessary gas chromatographic apparatus and establish
         operating parameters equivalent to those indicated in Table I.  By
         injecting calibration standards, establish the sensitivity limit of
         the detector and the linear range of the analytical system for each
         compound.
    6.3  Before using any cleanup procedure, the analyst must process a

-------
         series of calibration standards through the procedure to validate
         elution patterns and the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any samples, the analyst should demonstrate
         through the analysis of a distilled water method blank, that all
         glassware and reagents are interference-free.  Each time a set of
         samples is extracted or there is a change in reagents, a method
         blank should be processed as a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality assurance practices should be used with this
         method.  Field replicates should be collected to validate the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to'validate the accuracy of the
         analysis.  Where doubt exists over the identification of a peak on
         the chromatogram, confirmatory techniques such as mass spectroscopy
         should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prewashed with sample before collection.  Composite samples
         should be collected in refrigerated glass containers in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of
         collection until extraction.  Chemical preservatives should not be

-------
         used in the field unless more than 24 hours will elapse before
         delivery to the laboratory.  If the samples will not be extracted
         within 48 hours of collection, the sample should be adjusted to a
         pH range of 6.0-8.0 with sodium hydroxide or sulfuric acid.
    8.3  All samples must be extracted within 7 days and completely analyzed
         within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for later
         determination of sample volume.  Pour the entire sample into a
         two- liter separatory funnel.  Check the pH of the sample with
         wide-range pH paper and adjust to within the range of 5-9 with
         sodium hydroxide or sulfuric acid.
    9.2  Add 60 ml methyl ene chloride to the sample bottle, seal, and shake
         30 seconds to rinse the inner walls.  Transfer the solvent into the
         separatory funnel, and extract the sample by shaking the funnel for
         two minutes with periodic venting to release vapor pressure.  Allow
         the organic layer to separate from the water phase for a minimum of
         ten minutes.  If the emulsion interface between layers is more than
         one-third the size of the solvent layer, the analyst must employ
         mechanical techniques to complete the phase separation.  The
         optimum technique depends upon the sample, but may include
         stirring, filtration of the emulsion through glass wool, or
         centrifugation.  Collect the methylene chloride extract in a 250-ml
         Erlenmeyer flask.
    9.3  Add a second 60-ml volume of methylene chloride to the sample
         bottle and complete the extraction procedure a second time,
                             IS

-------
     combining the extracts in the Erlenmeyer flask.
9.4  Perform a third extraction in the same manner.  Pour the combined
     extract through a drying column containing 3-4 inches of anhydrous
     sodium sulfate, and collect it in a 500-ml Kuderna-Danish (K-D)
     flask equipped with a 10 ml concentrator tube.  Rinse the
     Erlenmeyer flask and column with 20-30 ml methylene chloride to
     complete the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus on a hot
     water bath (60-65°C) so that the concentrator tube is partially
     immersed in the hot water, and the entire lower rounded surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distillation
     the balls of the column will actively chatter but the chambers will
     not flood.  When the apparent volume of liquid reaches 1 ml, remove
     the K-D apparatus and allow it to drain for at least 10 minutes
     while cooling.  Remove the Snyder column and rinse the flask and
     its lower joint into the concentrator tube with 1-2 ml of methylene
     chloride.  A 5-ml syringe is recommended for this operation.
9.6  Add 1.0 ml toluene to the concentrator tube, and a clean boiling
     chip.  Attach a two-ball micro-Snyder column.  Prewet the
     micro-Snyder column by adding about 0.5 ml of methylene chloride to
     the top.  Place this micro-K-D apparatus on a water bath
     (60-65°C) so that the concentrator tube is partially immersed  in

                               134,

-------
         the hot water.   Adjust the vertical  position of the apparatus and
         water temperature as required to complete the concentration in 5 to
         10 minutes.   At the proper rate of distillation the balls will
         actively chatter but the chambers will  not flood.   When the
         apparent volume of liquid reaches 0.5 ml, remove the K-D apparatus
         and allow it to drain for at least 10 minutes while cooling.
         Remove the micro-Snyder column and rinse its lower joint into the
         concentrator tube with a small volume of toluene.   Adjust the final
         volume to 1.0 ml and stopper the concentrator tube and store
         refrigerated if further processing will not be performed
         immediately. Unless the sample is known to require cleanup,
         proceed with gas chromatographic analysis.
    9.7  Determine the original sample volume by refilling  the sample bottle
         to the mark and transferring the liquid to a 1000  ml graduated
         cylinder.  Record the sample volume to  the nearest 5 ml.
10.  Cleanup and Separation
    10.1 Prepare a slurry of lOg of activited Florisil in 10£ methylene
         chloride in hexane (V/V).  Use it to pack a 10 mm  ID chromatography
         column, gently  tapping the column to settle the Florisil.  Add 1 cm
         anhydrous sodium sulfate to the top of  the Florisil.
         10.1.2    Just  prior to exposure of the sodium sulfate layer to the
                   air transfer the 1 ml sample  extract onto the column
                   using an additional 2 ml of toluene to complete the
                   transfer.
         10.1.3    Just  prior to exposure of the sodium sulfate layer to the
                   air,  add 30 ml 10% methylene  chloride in hexane and

                              nr- 137

-------
                   continue the elution of the column.  Elution of the
                   column should be at a rate of about 2 ml per minute.
                   Discard the eluate from this fraction.
         10.1.4    Next elute the column with 30 ml  of 10% acetone/90%
                   methylene chloride (V/V) into a 500 ml K-D flask equipped
                   with a 10 ml concentrator tube.  Concentrate the
                   collected fraction by the K-D technique prescribed in 9.5
                   and 9.6, including the solvent exchange to 1 ml toluene.
                   This fraction should contain the nitroaromatics and
                   isophorone.
         10.1.5    Analyze by gas chromatography.
11.  Gas Chromatography
    11.1 Isophorone and nitrobenzene are analyzed by injection of a portion
         of the extract into a gas chromatograph with a flame ionization
         detector.   The dinitrotoluenes are analyzed by a separate injection
         into an electron capture gas chromatograph.  Table I summarizes
         some recommended gas chromatographic column materials and operating
         conditions for the instruments.  Included in this table are
         estimated  retention times and sensitivities that should be achieved
         by this method.  Examples of the separations achieved by the
         primary column are shown in Figures 1 and 2.  Calibrate the system
         daily with a minimum of three injections of calibration standards.
    11.2 Inject 2-5 ul of the sample extract using the solvent-flush
         technique.  Smaller (1.0 ul) volumes can be injected if automatic
         devices are employed.  Record the volume injected to the nearest
         0.05 ul, and the resulting peak size, in area units.

-------
    11.3 If the peak area exceeds the linear range of the  system, dilute  the
         extract and reanalyze.
    11.4 If the peak area measurement is prevented by the  presence of
         interferences, further cleanup is required.
12. Calculations
    12.1 Determine the concentration of individual compounds according to
         the formula:
         Concentration, ug/1 = (A) (B) (Vt)
                                 (Vi)  (Vs)
         where A = Calibration factor for chromatographic  system, in
                   nanograms material per area unit.
               B * Peak size in injection of sample extract, in area units
               V. = volume of extract injected (ul)
               V. = Volume of total extract (ul)
               V  = Volume of water extracted (ml)
    12.2 Report results in micrograms per liter without correction for
         recovery data.  When duplicate and spiked samples are analyzed,  all
         data obtained should be reported.
13. Accuracy and Precision
    The U.S. EPA Environmental Monitoring and Support Laboratory in
    Cincinnati is in the process of conducting an inter!aboratory method
    study to determine the accuracy and precision of this  test procedure.

                                 BIBLIOGRAPHY
"Development and Application of Test Procedures for Specific Organic Toxic
Substances in Wastewaters.  Category 4-Nitroaromatics and  Isophorone,"
Report for EPA Contract No.  68-03-2624 (In preparation).
                             jr -  13

-------
                                   TABLE I

             GAS CHROMATOGRAPHY OF .NITROAROMATICS AND ISOPHORONE

                            Retention Time (min.)      Detection Limit (ug/1)
Compound
Isophorone
Ni trobenzene
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Column 1
4.49
3.31
5.35
3.52
Column 2
5.72
4.31
6.54
4.75
EC
-
- •
0.06
0.06
FID
5
5
-
-
Column 1 conditions:  Gas-Chrom Q 80/100 mesh coated with 1.95% QF-1/1.5%
    OV-17 packed in a 4' x 1/4" OD glass column.  FID analysis for IP and NB
    requires nitrogen carrier gas at 44 ml/min and 85°C column
    temperature.  EC analysis for the DNTs requires 10% Methane/90% Argon
    carrier gas at 44 ml/min flow rate and 145°C column temperature.

Column 2 conditions:  Gas-Chrom Q 80/100 mesh coated with 3% OV-101 packed
    in a 10' x 1/4" OD glass column.  FID analysis of IP and NB requires
    nitrogen carrier gas at 44 ml/min flow rate and 100°C column
    temperature.  EC analysis for the DNTs requires 10% Methane/90% Argon
    carrier gas at 44 ml/min flow rate and 150°C column temperature.

Detection limit is calculated from the minimum detectable GC response being
    equal to five times the GC background noise, assuming a 10 ml final
    volume of the 1 liter sample extract, and assuming a GC injection of 5
    microliters.
                              JIT-  I MO

-------
    COLUMN: 1.5% OV-17+ 1.95% QF-1 ON GAS CHROM Q
  I  TEMPERATURE:  85°C.
    DETECTOR: FLAME IONIZAT10N
             LU
    J_
_L
02468
   RETENTION TIME-MINUTES

Figure 1.  Gas chromatogram of nitrobenzene and
           isophorone
          Hi-

-------
1.95°i
COLUMN:  1.5% OV-17
TEMPERATURE:  145°C.
DETECTOR: ELECTRON CAPTURE
                          QF-1 ON GAS CHROM Q
     2     46    8   10
   RETENTION TIME-MINUTES

Figure  2.  Gas chromatogram of dinitrotoluenes
             WO-

-------
                      POLYNUCLEAR AROMATIC HYDROCARBONS
                                  METHOD 610

1.   Scope and Application
    1.1  This method covers the determination of certain polynuclear
         aromatic hydrocarbons (PAH).  The following parameters may be
         determined by this method:
         Parameter           STORET No.      Parameter             STORET No.
         Acenaphthene           34205        Chrysene                 34320
         Acenaphthylene         34200        Dibenzo(ah-)anthracene    34556
         Anthracene             34220        Fluoranthene             34376
         Benzo(a)anthracene     34526        Fluorene                 34381
         Benzo(a)pyrene         34247        Indeno(l,2,3-cd)pyrene   34403
         Benzo(b)fluoranthene   34230        Naphthalene              34696
         Benzo(ghi)perylene     34521        Phenanthrene             34461
         Benzo(k)fluoranthene   34242        Pyrene                   34469
    1.2  This method is applicable to the determination of these compounds
         in municipal and industrial discharges.  It is designed to be used
         to meet the monitoring requirements of the National Pollutant
         Discharge Elimination System (NPDES).  As such, it presupposes a
         high expectation of finding the specific compounds of interest.  If
         the user is attempting to screen samples for any or all of the
         compounds above, he must develop independent protocols for the
         verification of identity.
    1.3  This method contains both liquid and gas chromatographic
         approaches, depending upon the needs of the analyst.  The gas

-------
         chromatographic procedure cannot adequately resolve the following
         four pairs of compounds:   anthracene and phenanthrene; chrysene and
         benzo(a)anthracene;  benzo(b)fluoranthene and benzo(k)fluoranthene;
         and dibenzo(ah)anthracene and indeno(l,2,3-cd)pyrene.  Unless the
         purposes  of the analysis  can be served by reporting a sum for an
         unresolved pair, the liquid chromatographic approach must be used
         for these compounds.  The liquid chromatographic method will
         resolve all of the 16 compounds listed above.
    1.4  The sensitivity of this method is usually dependent upon the level
         of interferences rather than instrumental limitations.  The limits
         of detection listed  in Table I for the liquid chromatographic
         approach  represent sensitivities that can be achieved in
         wastewaters in the absence of interferences.
    1.5  This method is recommended for use only by experienced residue
         analysts  or under the close supervision of such qualified persons.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
         using separatory funnel techniques.  The extract is dried and
         concentrated to a volume of 10 ml or less.  Chromatographic
         conditions are described which allow for the accurate measurement
         of the compounds in the extract by either High Performance Liquid
         Chromatography (HPLC) or gas chromatography.
    2.2  If interferences are encountered, the method provides a selected
         general purpose cleanup procedure to aid the analyst  in their
         elimination.
3.  Interferences
                               77T

-------
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of the chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
         in all-glass systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the
         industrial complex or municipality being sampled.  While a general
         clean-up technique is provided as part of this method, unique
         samples may require additional clean-up approaches to achieve the
         sensitivities stated in Table 1.
    3.3  The extent of interferences that may be encountered using liquid
         chromatographic techniques has not been fully assessed.  Although
         the chromatographic conditions described allow for a unique
         resolution of the specific PAH compounds covered by this method,
         other PAH compounds may interfere.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1  Grab sample bottle - amber glass, 1-liter or 1-quart
                volume.  French or Boston Round design is recommended.  The
                container must be washed and solvent rinsed before use to
                minimize interferences.
         4.1.2  Bottle caps - Threaded to screw on to the sample bottles.
                Caps must be lined with Teflon.  Foil may be substituted if

-------
            sample is not corrosive.
     4.1.3  Compositing equipment - Automatic or manual compositing
            system. Must incorporate glass sample containers for the
            collection of a minimum of 250 ml.  Sample containers must
            be kept refrigerated during sampling.  No tygon or rubber
            tubing may be used in the system.
4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
     frit.
4.4  Kuderna-Danish (K-D) Apparatus
     4.4.1  Concentrator tube - 10 ml, graduated (Kontes K-570050-1025
            or equivalent).  Calibration must be checked.  Ground glass
            stopper (size 19/22 joint) is used to prevent evaporation of
            extracts.
     4.4.2  Evaporative flask - 500 ml (Kontes K-57001-0500 or
            equivalent).  Attach to concentrator tube with springs.
            (Kontes K-662750-0012).
     4.4.3  Snyder column - three-ball macro  (Kontes K503000-0121 or
            equivalent).
     4.4.4  Snyder column - two-ball micro (Kontes K-569001-0219 or
            equivalent).
     4.4.5  Boiling chips - solvent extracted, approximately 10/40 mesh.
4.5  Water bath - Heated, with concentric ring cover, capable of
     temperature control (_+ 2°C).  The bath  should be used in a hood.
4.6  HPLC Apparatus:
     4.6.1  Gradient pumping system, constant flow.

-------
         4.6.2  Reverse phase column,  HC-ODS Sil-X, 250 mm x 2.6 mm ID
                (Perkin Elmer No.  809-0716 or equivalent).
         4.6.3  Fluorescence detector, for excitation at 280 ran and emission
                at 389 nm.
         4.6.4  UV detector, 254 nm,  coupled to fluorescence detector.
         4.6.5  Strip chart recorder  compatible with detectors, (A data
                system for  measuring  peak areas is recommended).
    4.7  Gas chromatograph  - Analytical system complete with gas
         chromatograph suitable for on-column injection and all required
         accessories including dual flame ionization detectors, column
         supplies, recorder, gases, syringes.  A data system for measuring
         peak areas is recommended.
    4.8  Chromatographic column - 250 mm long x 10 mm ID with coarse fritted
         •disc at bottom and Teflon stopcock.
5.  Reagents
    5.1  Preservatives:
         5.1.1  Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2  Sulfuric acid - (ACS)  Mix equal volumes of cone. H2$04
                with distilled water.
         5.1.3  Sodium thiosulfate -  (ACS) Granular.
    5.2  Methylene chloride, Pentane,  Cyclohexane, High Purity Water-HPLC
         quality, distilled in glass.
    5.3  Sodium sulfate - (ACS) Granular, anhydrous (purified by heating at
         400°C for 4 hrs. in a shallow tray).
    5.4  Stock standards -  Prepare stock standard solutions at a
         concentration of 1.00 ug/ul  by dissolving 0.100 grams of assayed

-------
         reference material In pesticide quality isooctane or other
         appropriate solvent and diluting to volume in a 100 ml ground glass
         stoppered volumetric flask.  The stock solution is transferred to
         ground glass stoppered reagent bottles, stored in a refrigerator,
         and checked frequently for signs of degradation or evaporation,
         especially just prior to preparing working standards from them.
    5.5  Acetonitrile - Spectral quality.
    5.6  Silica gel - 100/120 mesh desiccant (Davison Chemical grade 923 or
         equivalent). 'Before use, activate for at least 16 hours at 130°C
         in a foil covered glass container.
6.  Calibration
    6.1  Prepare calibration standards that contain the compounds of
         interest, either singly or mixed together.  The standards should be
         prepared at concentrations covering two or more orders of magnitude
         that will completely bracket the working range of the
         chromatographic system.  If the sensitivity of the detection system
         can be calculated from Table I as 100 ug/1 in the final extract,
         for example, prepare standards at 10 ug/1, 50 ug/1, 100 ug/1, 500
         ug/1, etc. so that injections of 1-5 ul of each calibration
         standard will define the linearity of the detector in the working
         range.
    6.2  Assemble the necessary HPLC or gas chromatographic apparatus and
         establish operating parameters equivalent to those indicated in
         Table I or II.  By injecting calibration standards, establish the
         sensitivity limit of the detectors and the linear range of the
         analytical systems for each compound.

-------
    6.3  Before using any cleanup procedure, the analyst must process a
         series of calibration standards through the procedure to validate
         elution patterns and the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any samples, the analyst should demonstrate
         through the analysis of a distilled water method blank, that all
         glassware and reagents are interference-free.  Each time a set of
         samples is extracted or there is a change in reagents, a method
         blank should be processed as a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality assurance practices should be used with this
         method.  Field replicates should be collected to validate the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification of a peak on
         the chromatogram, confirmatory techniques such as fraction
         collection and GC-mass spectroscopy should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prewashed with sample before collection.  Composite samples
         should be collected in refrigerated glass containers in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of

-------
         collection until extraction.  Chemical preservatives should not be
         used in the field unless more than 24 hours will elapse before
         delivery to the laboratory.  If the samples will not be extracted
         within 48 hours of collection, adjust the sample to a pH range of
         6.0-8.0 with sodium hydroxide or sulfuric acid and add 35 mg sodium
         thiosulfate per part per million of free chlorine per liter.
    8.3  All samples must be extracted within 7 days and completely analyzed
         within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for later
         determination of sample volume.  Pour the entire sample into a
         two-liter separatory funnel.  Check the pH of the sample with
         wide-range pH paper and adjust to within the range of 5-9 with
         sodium hydroxide or sulfuric acid.
    9.2  Add 60 ml methylene chloride to the sample bottle, seal, and shake
         30 seconds to rinse the inner walls.  Transfer the solvent into the
         separatory funnel, and extract the sample by shaking the funnel for
         two minutes with periodic venting to release vapor pressure.  Allow
         the organic layer to separate from the water phase for a minimum of
         ten minutes.  If the emulsion interface between layers is more than
         one-third the size of the solvent layer, the analyst must employ
         mechanical techniques to complete the phase separation.  The
         optimum technique depends upon the sample, but may include
         stirring, filtration of the emulsion through glass wool, or
         centrifugation.  Collect the methylene chloride extract in a 250-ml
         Erlenmeyer flask.

                            JIT- 148

-------
9.3  Add a second 60-ml volume of methylene chloride to the sample
     bottle and complete the extraction procedure a second time,
     combining the extracts in the Erlenmeyer flask.
9.4  Perform a third extraction in the same manner.  Pour the combined
     extract through a drying column containing 3-4 inches of anhydrous
     sodium sulfate, and collect it in a 500-ml Kuderna-Danish (K-D)
     flask equipped with a 10 ml concentrator tube.  Rinse the
     Erlenmeyer flask and column with 20-30 ml methylene chloride to
     complete the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus on a hot
     water bath (60-65°C) so that the concentrator tube is partially
     immersed in the hot water, and the entire lower rounded surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distillation
     the balls of the column will  actively chatter but the chambers will
     not flood.  When the apparent volume of liquid reaches 1  ml, remove
     the K-D apparatus and allow it to drain for at least 10 minutes
     while cooling.  Remove the Snyder column and rinse the flask and
     its lower joint into the concentrator tube with 1-2 ml of methylene
     chloride.  A 5-ml syringe is  recommended for this operation.
     Stopper the concentrator tube and store refrigerated if further
     processing will not be performed immediately.
9.6  Determine the original  sample volume by refilling the sample bottle

-------
         to the mark and transferring the liquid to a 1000 ml graduated
         cylinder.   Record the sample volume to the nearest 5 ml.
    9.7  If the sample requires cleanup before chromatographlc analysis,
         proceed to Section 10.  If the sample does not require cleanup, or
         if the need for cleanup is unknown, analyze an aliquot of the
         extract according to Section 11 or Section 12.
10.  Cleanup and Separation
    10.1 Before the silica gel cleanup technique can be utilized, the
         extract solvent must be exchanged to cyclohexane.  Add a 1-10 ml
         aliquot of sample extract (in methylene chloride) and a boiling
         chip to a clean K-0 concentrator tube.  Add 4 ml cyclohexane and
         attach a micro-Snyder column.  Prewet the micro-Snyder column by
         adding 0.5 ml methylene chloride to the top.  Place the micro-K-D
         apparatus on a boiling (100°C) water bath so that the
         concentrator tube is partially immersed in the hot water.  Adjust
         the vertical position of the apparatus and the water temperature as
         required to complete concentration in 5-10 minutes.  At the proper
         rate of distillation the balls of the column will actively chatter
         but the chambers will not flood.  When the apparent volume of the
         liquid reaches 0.5 ml, remove the K-D apparatus  and allow .it to
         drain for at least 10 minutes while cooling.  Remove the
         micro-Snyder column and rinse its lower joint into the concentrator
         tube with a minimum of cyclohexane.  Adjust the  extract volume to
         about 2 ml.
    10.2 Silica Gel Column Cleanup for PAHs.
         10.2.1 Prepare a slurry of lOg activated silica  gel in methylene

-------
                chloride and place this in a 10 mm ID chromatography
                column.  Gently tap the column to settle the silica gel and
                elute the methylene chloride.  Add 1-2 cm of anhydrous
                sodium sulfate to the top of the silica gel.
         10.2.2 Preelute the column with 40 ml pentane.  Discard the eluate
                and just prior to exposure of the sodium sulfate layer to
                the air, transfer the 2 ml cyclohexane sample extract onto
                the column, using an additional 2 ml of cyclohexane to
                complete the transfer.
         10.2.3 Just prior to exposure of the sodium sulfate layer to the
                air, add 25 ml pentane and continue elution of the column.
                Discard the pentane eluate.
         10.2.4 Elute the column with 25 ml of 40% methylene chloride/60%
                pentane and collect the eluate in a 500 ml K-D flask
                equipped with a 10 ml concentrator tube.  Elution of the
                column should be at a rate of about 2 ml/min.
         10.2.5 Concentrate the collected fraction to less than 10 ml by K-D
                techniques, as in 9.5, using pentane to rinse the walls of
                the glassware.  Proceed with HPLC or gas chromatographic
                analysis.
11. High Performance Liquid Chromatography HPLC
    11.1 To the extract in the concentrator tube, add 4 ml acetonitrile and
         a new boiling chip, then attach a micro-Snyder column.  Increase
         the temperature of the hot water bath to 95-100°C.  Concentrate
         the solvent as above.  After cooling, remove the micro-Snyder
         column and rinse its lower joint into the concentrator tube with

-------
         about 0.2 ml  acetonitrile.   Adjust  the  extract  volume to 1.0 ml.
    11.2  Table I  summarizes  the  recommended  HPLC column  materials and
         operating conditions  for  the instrument.   Included in this  table
         are  estimated retention times and  sensitivities that should be
         achieved by this method.  An example  of the  separation achieved by
         this column is shown  in Figure 1.   Calibrate the system daily with
         a minimum of  three  injections of calibration standards.
    11.3  Inject  2-5 ul of the  sample  extract with a high pressure syringe  or
         sample  injection loop.  Record the  volume injected to the nearest
         0.05 ul, and  the resulting peak size,  in area units.
    11.4  If the  peak area exceeds  the linear range of the system, dilute the
         extract and reanalyze.
    11.5  If the  peak area measurement is prevented by the presence of
         interferences, further  cleanup is  required.
    11.6  The  UV  detector is  recommended for the  determination of naphthalene
         and  acenaphthylene  and  the fluorescence detector is recommended for
         the  remaining PAHs.
12.  Gas Chromatography
    12.1  The  gas chromatographic procedure  will  not resolve certain isomeric
         pairs as indicated  in Table  II. The  liquid  chromatographic
         procedure  (Section  11)  must  be used for these materials.
    12.2  To  achieve maximum  sensitivity with this method, the extract must
         be  concentrated to  1.0 ml.   Add a  clean boiling chip to the
         methylene  chloride  extract  in the  concentrator tube.  Attach a
         two-ball micro-Snyder column.  Prewet the micro-Snyder column by
         adding  about  0.5 ml of methylene chloride to the top.  Place this

                                KT

-------
         micro-K-D apparatus  on  a hot  water  bath (60-65°C)  so that the
         concentrator  tube  is partially immersed in  the  hot water.  Adjust
         the  vertical  position of the  apparatus and  water temperature as
         required  to complete the concentration in 5 to  10 minutes.   At the
         proper rate of  distillation the balls  will  actively chatter but the
         chambers  will not  flood.  When the  apparent volume of liquid
         reaches 0.5 ml,  remove  the K-D apparatus and allow it to drain for
         at  least  10 minutes  while cooling.   Remove  the  micro-Snyder column
         and  rinse its lower  joint into the  concentrator tube with a small
         volume of methylene  chloride.   Adjust  the final  volume to 1.0 ml
         and  stopper the  concentrator  tube.
    12.3 Table II  describes the  recommended  gas chromatographic column
         material  and  operating  conditions for  the  instrument.  Included in
         this table are  estimated retention  times that should be achieved by
         this method.  Calibrate the gas chromatographic system daily with a
         minimum of three injections of calibration  standards.
    12.4 Inject 2-5 ul of the sample extract using the solvent-flush
         technique. Smaller  (1.0 ul)  volumes can be injected if automatic
         devices are employed.  Record the volume injected to the nearest
         0.05 ul,  and  the resulting peak size,  in area units.
    12.5 If  the peak area exceeds the  linear range of the system, dilute the
         extract and reanalyze.
    12.6 If  the peak area measurement  is prevented by the presence of
         interferences,  further  cleanup is required.
13.  Calculations
    13.1 Determine the concentration of individual compounds according to

                                TV-

-------
         the formula:
         Concentration,  ug/1  - (A) (B) (Vt)
                                 (Vi)   (Vs)
         where A = Calibration factor  for chromatographic system, in
                   nanograms  material  per area unit.
               B = Peak  size  in injection of sample extract, in area units
               V.  = volume of extract  injected (ul)
               V.  = Volume of total extract (ul)
               V  = Volume of water extracted (ml)
    13.2 Report results  in micrograms  per liter without correction for
         recovery data.   When duplicate and spiked samples are analyzed, all
         data obtained should be reported.
14.  Accuracy and Precision
    14.1 The U.S.  EPA Environmental Monitoring and Support Laboratory in
         Cincinnati is in the process  of conducting an inter!aboratory
         method study to determine the accuracy and precision of this test
         procedure.
                                 BIBLIOGRAPHY
"Development and Application  of Test Procedures for Specific Organic Toxic
Substances in Wastewaters, Category 9-PAHs."  Report for EPA Contract
68-03-2624 (In preparation).
                               '13.

-------
                                   TABLE I

               High Performance Liquid Chromatography of PAH's


                                                      Detection Limit (ug/1)
Compound                   Retention Time (min)         UV	Fluorescence
Naphthalene
Acenaphthylene
Acenaphthene
Fluorene
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benzo( a) anthracene
Chrysene
Benzo (b )f 1 uorant hene
Benzo(k)fluoranthene
Benzo (a)pyrene
Dibenzo(a,h) anthracene
Benzo (ghi)perylene
Indenof 1 ,2,3-cd)pyrene
16.17
18.10
20.14
20.89
22.32
23.78
25.00
25.94
29.26
30.14
32.44
33.91
34.95
37.06
37.82
39.21
2.5
5.0
3.0
0.5
0.25
0.10
0.50
0.10
0.20
0.20
1.0
0.30
0.25
1.0
0.75
0,30
20.0
100.0
4.0
2.0
1.2
1.5
0.05
0.05
0.04
0.5
0.04
0.04
0.04
0.08
0.2
0.1
HPLC conditions:  Reverse phase HC-ODS Sil-X 2.6 x 250 mm Perkin-Elmer
    column; isocratic elution for 5 min. using 40% acetonitrile/60% water,
    then linear gradient elution to 100% acetonitrile over 25 minutes; flow
    rate is 0.5 ml/min.

 Detection limit is calculated from the minimum detectable HPLC response
    being equal to five times the background noise, assuming an equivalent
    of a 2 ml final volume of the 1 liter sample extract, and assuming an
    HPLC injection of 2 microliters.

-------
                                   TABLE II

                          Gas Chromatography of PAHs

Compound                            Retention Time (Min)

Naphthalene                                 4.5
Acenaphthylene                             10.4
Acenaphthene                               10.8
Fluorene                                   12.6
Phenanthrene                               15.9
Anthracene                                 15.9
Fluoranthene                               19.8
Pyrene                                     20.6
Benzo(a)anthracene                         20.6
Chrysene                                   24.7
Benzo(b)fluoranthene                       28.0
Benzo(k)fluoranthene                       28.0
Benzo(a)pyrene                             29.4
Dibenzo(a,h)anthracene                     36.2
Indeno(l,2,3-cd)pyrene                     36.2
Benzo(ghi)pery!ene	38.6	


6C conditions:  Chromosorb W-AW-DCMS 100/120 mesh coated with 3% OV-17,
packed in a 61 x 2 mm ID glass column, with nitrogen carrier gas at 40
ml/min flow rate.  Column temperature was held at 100°C for 4 minutes,
then programmed at 8°/minute to a final hold at 280°C.

-------
COLUMN:  HC-ODS S1L-X
MOBILE PHASE: 40% TO 100% ACETONITRILE IN WATER
DETECTOR: FLUORESCENCE
                                  28
36
40
                    RETENTION TIME-MINUTES

 Figure 1.  Liquid chromatogram of polynuclear aromatics

-------
                                  HALOETHERS
                                  METHOD 611

1.   Scope and Application
    1.1  This method covers the determination of certain haloethers.  The
         following parameters may be determined by this method:
         Parameter                                        STORET No.
         Bis(2-chloroethyl) ether                          34273
         Bis(2-chloroethoxy) methane                       34278
         Bis(2-chloroisopropyl) ether                      34283
         4-Bromophenyl phenyl ether                        34636
         4-Chlorophenyl phenyl ether                       34641

    1.2  This method is applicable to the determination of these compounds
         in municipal and industrial discharges.  It is designed to be used
         to meet the monitoring requirements of the National Pollutant
         Discharge Elimination System (NPDES).  As such, it presupposes a
         high expectation of finding the specific compounds of interest.  If
         the user is attempting to screen samples for any or all of the
         compounds above, he must develop independent protocols for the
         verification of identity.
    1.3  The sensitivity of this method is usually dependent upon the level
         of interferences rather than instrumental limitations.  The limits
         of detection listed in Table I represent sensitivities that can be
         achieved in wastewaters in the absence of interferences.
    1.4  This method is recommended for use only by experienced residue

-------
         analysts or under the close supervision of such qualified persons.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
         using separatory funnel techniques.  The extract is dried and
         concentrated to a volume of 10 ml or less.  Chromatographic
         conditions utilizing a halide specific detector are described which
         allow for the accurate measurement of the compounds in the extract.
    2.2  If interferences are encountered, the method provides a selected
         general purpose cleanup procedure to aid the analyst in their
         elimination.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
         in all-glass systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the
         industrial complex or municipality being sampled.  While general
         clean-up techniques are provided as part of this method, unique
         samples may require additional cleanup approaches to achieve the
         sensitivities stated in Table I.
    3.3  Dichlorobenzenes are known to coelute with haloethers under some

-------
         gas chromatographic conditions.  If these materials are present
         together in a sample,  it may be necessary to analyze the extract
         with two different column packings to completely resolve all of the
         compounds.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1  Grab sample bottle - amber glass, 1-liter or 1-quart
                volume.  French or Boston Round design is recommended.  The
                container must be washed and solvent rinsed before use to
                minimize interferences.
         4.1.2  Bottle caps - Threaded to screw on to the sample bottles.
                Caps must be lined with Teflon.  Foil may be substituted if
                sample is not corrosive.
         4.1.3  Compositing equipment - Automatic or manual compositing
                system. Must incorporate glass sample containers for the
                collection of a minimum of 250 ml.  Sample containers must
                be kept refrigerated during sampling.  No tygon or rubber
                tubing may be used in the system.
    4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
    4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
         frit.
    4.4  Kuderna-Danish (K-D) Apparatus
         4.4.1  Concentrator tube - 10 ml, graduated (Kontes K-570050-1025
                or equivalent).  Calibration must be checked.  Ground glass
                stopper (size 19/22 joint) is used to prevent evaporation of
                extracts.
                              Tf-

-------
         4.4.2  Evaporative flask - 500 ml (Kontes K-57001-0500 or
                equivalent).  Attach to concentrator tube with springs.
                (Kontes K-662750-0012).
         4.4.3  Snyder column - three-ball macro (Kontes K503000-0121 or
                equivalent).
         4.4.4  Snyder column - two-ball micro (Kontes K-569001-0219 or
                equivalent).
         4.4.5  Boiling chips - solvent extracted, approximately 10/40 mesh.
    4.5  Water bath - Heated, with concentric ring cover, capable of
         temperature control (+; 2°C).  The bath should be used in a hood.
    4.6  Gas chromatograph - Analytical system complete with gas chromato-
         graph suitable for on-column injection and all required accessories
         including halide specific detector, column supplies, recorder,
         gases, syringes.  A data system for measuring peak areas is
         recommended.
    4.7  Chromatographic Column - 400 mm long x 19 mm ID with coarse fritted
         plate on bottom and Teflon stopcock (Kontes K-420540-0224 or
         equivalent).
5.  Reagents
    5.1  Preservatives:
         5.1.1  Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2  Sulfuric acid (1+1) - (ACS) Mix equal volumes of cone.
                H2S04 with distilled water.
    5.2  Methylene chloride - Pesticide quality or equivalent.
    5.3  Sodium Sulfate - (ACS) Granular, anhydrous (purified by heating at
         400°C for 4 hrs. in a shallow tray).

-------
    5.4  Stock standards - Prepare stock standard solutions at a concentra-
         tion of 1.00 ug/ul  by dissolving 0.100 grams of assayed reference
         material  in pesticide quality acetone or other appropriate solvent
         and diluting to volume in a 100 ml  ground glass stoppered volu-
         metric flask.  The stock solution is transferred to ground glass
         stoppered reagent bottles,  stored in a refrigerator, and checked
         frequently for signs of degradation or evaporation, especially just
         prior to  preparing working  standards from them.
    5.5  Florisil  - PR Grade (60/100 mesh);  purchase activated at 1250°F
         and store in the dark in glass containers with glass stoppers or
         foil-lined screw caps.  Before use, activate each batch overnight
         at 130°C  in a foil-covered  glass container.
    5.6  Hexane, Petroleum ether (boiling range 30-60°C) - pesticide
         quality or equivalent.
    5.7  Diethyl Ether - Nanograde,  redistilled in glass, if necessary.
         5.7.1  Must be free of peroxides as indicated by EM Quant test
                strips.  (Test strips are available from EM Laboratories,
                Inc., 500 Executive Blvd., Elmsford, N.Y.  10523.)
         5.7.2  Procedures recommended for removal of peroxides are provided
                with the test strips.  After cleanup 20 ml ethyl alcohol
                preservative must be added to each liter of ether.
6.  Calibration
    6.1  Prepare calibration standards that contain the compounds of
         interest, either singly or mixed together.  The standards should be
         prepared at concentrations covering two or more orders of magnitude
         that will completely bracket the working range of the chromato-

-------
         graphic system.  If the sensitivity of the detection system can be
         calculated from Table I as 100 ug/1 in the final extract, for
         example, prepare standards at 10 ug/1, 50 ug/1, 100 ug/1, 500 ug/1,
         etc. so that injections of 1-5 ul of each calibration standard will
         define the linearity of the detector in the working range.
    6.2  Assemble the necessary gas chromatographic apparatus and establish
         operating parameters equivalent to those indicated in Table I.  By
         injecting calibration standards, establish the sensitivity limit of
         the detector and the linear range of the analytical system for each
         compound.
    6.3  The cleanup procedure in Section 10 utilizes Florisil chromatog-
         raphy.  Florisil from different batches or sources may vary in
         absorption capacity.  To standardize the amount of Florisil which
         is used, the use of lauric acid value (Mills, 1968) is suggested.
         The referenced procedure determines the adsorption from hexane
         solution of lauric acid (mg)  per gram Florisil.  The amount of
         Florisil to be used for each  column is calculated by dividing 110
         by this ratio and multiplying by 20 grams.
    6.4  Before using any cleanup procedure, the analyst must process a
         series of calibration standards through the procedure to validate
         elution patterns and the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any samples, the analyst should demonstrate
         through the analysis of a distilled water method blank, that all
         glassware and reagents are interference-free.  Each time a set of
         samples is extracted or there is a change in reagents, a method

-------
         blank should be processed as a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality assurance practices should be used with this
         method.   Field replicates should be collected to validate the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to validate the accuracy of the
         analysis.  Where doubt exists over the identification of a peak on
         the chromatogram, confirmatory techniques such as mass spectroscopy
         should be used.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prewashed with sample before collection.  Composite samples
         should be collected in refrigerated glass containers in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of collection
         until extraction.  Chemical preservatives should not be used in the
         field unless more than 24 hours will elapse before delivery to the
         laboratory.  If the samples will not be extracted within 48 hours
         of collection, the sample should be adjusted to a pH range of
         6.0-8.0 with sodium hydroxide or sulfuric acid.
    8.3  All samples must be extracted within 7 days and completely analyzed
         within 30 days of collection.

-------
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for later
         determination of sample volume.  Pour the entire sample into a
         two-liter separatory funnel.  Check the pH of the sample with
         wide-range pH paper and adjust to within the range of 5-9 with
         sodium hydroxide or sulfuric acid.
    9.2  Add 60 ml methylene chloride to the sample bottle, seal, and shake
         30 seconds to rinse the inner walls.  Transfer the solvent into the
         separatory funnel, and extract the sample by shaking the funnel for
         two minutes with periodic venting to release vapor pressure.  Allow
         the organic layer to separate from the water phase for a minimum of
         ten minutes.  If the emulsion interface between layers is more than
         one-third the size of the solvent layer, the analyst must employ
         mechanical techniques to complete the phase separation.  The
         optimum technique depends upon the sample, but may include stir-
         ring, filtration of the emulsion through glass wool, or centrifu-
         gation.  Collect the methylene chloride extract in a 250-ml
         Erlenmeyer flask.
    9.3  Add a second 60-ml volume of methylene chloride to the sample
         bottle and complete the extraction procedure a second time,
         combining the extracts in the Erlenmeyer flask.
    9.4  Perform a third extraction in the same manner.  Pour the combined
         extract through a drying column containing 3-4 inches of anhydrous
         sodium sulfate, and collect it in a 500-ml Kuderna-Danish (K-D)
         flask equipped with a 10 ml concentrator tube.  Rinse the
         Erlenmeyer flask and column with 20-30 ml methylene chloride to

-------
     complete the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus on a hot
     water bath (60-65°C) so that the concentrator tube rs partially
     immersed in the hot water, and the entire lower rounded surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distillation
     the balls of the column will actively chatter but the chambers will
     not flood.  When the apparent volume of liquid reaches 1-2  ml,
     remove the K-D apparatus and allow it to drain for at least 10
     minutes while cooling.  NOTE:  Haloethers have a sufficiently high
     volatility that significant losses will occur in concentration
     steps if care is not exercised.  It is important to maintain a
     constant gentle evaporation rate and not to allow the liquid volume
     to fall below 1-2 ml before removing the K-0 from the hot water
     bath.
9.6  Momentarily remove the Snyder column, add 50 ml hexane and a new
     boiling chip and replace the column.  Raise the temperature of the
     water bath to 85-90°C.  Concentrate the extract as in 9.5 except
     use hexane to prewet the column.  Remove the Snyder column and
     rinse the flask and its lower joint into the concentrator tube with
     1-2 ml hexane.  Stopper the concentrator tube and store refrige-
     rated if further processing will not be performed immediately.
9.7  Determine the original sample volume by refilling the sample bottle

-------
         to the mark and transferring the liquid to a 1000 ml graduated
         cylinder.   Record the sample volume to the nearest 5 ml.
    9.8  Unless the sample is known to require cleanup, proceed to analysis
         by gas chromatography.
10.  Cleanup and Separation
    10.1 Florisil  Column Cleanup for Haloethers
         10.1.1 Adjust the sample extract volume to 10 ml.
         10.1.2 Place a charge (nominally 20 g but determined in Section
                6.3) of activated Florisil in a 19 mm ID chromatography
                column.  After settling the Florisil  by tapping the column,
                add about one-half inch layer of anhydrous granular sodium
                sulfate to the top.
         10.1.3 Pre-elute the column, after cooling,  with 50-60 ml of
                petroleum ether.  Discard the eluate  and just prior to
                exposure of the sulfate layer to air, quantitatively
                transfer the sample extract into the  column by decantation
                and subsequent petroleum ether washings.  Discard the
                eluate.  Just prior to exposure of the sodium sulfate layer
                to the air, begin eluting the column  with 300 ml of 6% ethyl
                ether/94% petroleum ether.  Adjust the elution rate to
                approximately 5 ml/min and collect the eluate in a 500 ml
                K-D flask equipped with a 10 ml concentrator tube.  This
                fraction should contain all of the haloethers.
         10.1.4 Concentrate the fraction by K-D as in 9.5 except prewet the
                Snyder column with hexane.  When the  apparatus is cool,
                remove the column and rinse the flask and its lower joint
                                 f-  Itfc

-------
                into the concentrator tube with 1-2 ml hexane.  Analyze by
                gas chromatography.
11.  Gas Chromatography
    11.1 Table I summarizes some recommended gas chromatographic column
         materials and operating conditions for the instrument.  Included in
         this table are estimated retention times and sensitivities that
         should be achieved by this  method.  Examples of the separations
         achieved by these columns are shown in Figures 1 and 2.  Calibrate
         the system daily with a minimum of three injections of calibration
         standards.
    11.2 Inject 2-5 ul of the sample extract using the solvent-flush
         technique.  Smaller (1.0 ul) volumes can be injected if automatic
         devices are employed.  Record the volume injected to the nearest
         0.05 ul, and the resulting  peak size,  in area units.
    11.3 If the peak area exceeds the linear range of the system, dilute the
         extract and reanalyze.
    11.4 If the peak area measurement is prevented by the presence of inter-
         ferences, further cleanup is required.
12.  Calculations
    12.1 Determine the concentration of individual compounds according to
         the formula:
         Concentration, ug/1 = (A) (B) (Vt)
                                 (Vi)  (Vs)
         where A = Calibration factor for chromatographic system, in nano-
                   grams material per area unit.
                               UT-

-------
               B * Peak size in injection of sample extract, in area units
               "V.. = volume of extract injected (ul)
               V. = Volume of total extract (ul)
               V  » Volume of water extracted (ml)
    12.2 Report results in micrograms per liter without correction for
         recovery data.  When duplicate and spiked samples are analyzed, all
         data" obtained should be reported.
13. Accuracy and Precision
    The U.S. EPA Environmental Monitoring and Support Laboratory in
    Cincinnati is in the process of conducting an interlaboratory method
    study to determine the accuracy and precision of this test procedure.

                                 BIBLIOGRAPHY
1.  "Development and Application of Test Procedures for Specific Organic
    Toxic Substances in Wastewaters.  Category 2-Haloethers."  Report for
    EPA Contract 68-03-2633 (In preparation).

2.  Mills, P.A., "Variation of Florisil  Activity:  Simple Method for
    Measuring Absorbent Capacity and Its Use in Standardizing Florisil
    Columns," Journal  of the Association of Official Analytical Chemists,
    51, 29 (1968).

-------
                                   TABLE I
                       Gas Chromatography of Haloethers
Compound
Retention Time  (min)

 Column 1   Column 2
 Detection

Limit (ug/L)
Bis(2-chlorpisopropyl) ether
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy) methane
4-Chlorophenyl phenyl ether
4-Bromophenyl phenyl ether
8.41
9.32
13.1
19.4
21.2
9.70
9.06
9.97
15.0
16.2
0.9
0.5
0.4
2.2
1.1
Column 1 conditions:  Supelcoport 100/120 mesh coated with 3% SP-1000 packed
    in 1.8 m long x 2.1 mm ID glass column with ultra-high purity helium
    carrier/gas at 40 ml/min flow rate.  Column temperature is 60°C for 2
    minutes after injection then program at 8°C/min to 230°C and hold
    for 4 minutes.  Under these conditions R.T. of Aldrin is 22.6 minutes.

Column 2 conditions:  Tenax-GC 60/80 mesh packed in a 1.8 m long x 2.1 mm ID
    glass column with helium carrier gas at 40 ml/min flow rate.  Column
    temperature 150°C for 4 minutes after injection then program at
    16°C/min to 310°C.  Under these conditions R.T. of Aldrin is 18.4
    minutes.

 Detection limit is calculated from the minimum detectable GC response being
    equal to five times the GC background noise, assuming a 10 ml final
    volume of the 1 liter sample extract, and assuming a GC injection of 5
    micro!iters.  These values were collected using the Tracor 700 Hall
    electrolytic conductivity detector with furnace temperature 900°C,
    transfer line 250°C, 95% ethanol electrolyte at 0.3 ml/min flow rate,
    and hydrogen reaction gas at 60 ml/min.
                              HE-It 9

-------
COLUMN:  3!S SP-1000 ON SUPELCOPORT
PROGRAM: 60°C-2 MINUTES 8"/MINUTE TO  230°C.
DETECTOR: HALL ELECTROLYTIC CONDUCTIVITY
cc
III


ce
LU
X
LU
O_
25
LU
z
X
LU
X






Q.
O
o
cc
O
X
CJ
t
Psl
—
m
•
1
fl
I K
**•
X
LU
Ij
X
LU
O
cc
o
X
CJ
1
CM
CO
CO





^ 	
1 1 1 1 1
} 2 ^ 6 8 10




LU

X
LU
&

X
O
X
LU
i
o
X
CJ
fsi
to
CO
JL_

12 14






ce
LU
X
h-
LU
_J
LU
Q.
-J
Z»
LU
Q.
O
ce
O

CJ
_JU

16 18


ce
X ^
LU cc
-i Q
Z <
LU
X
CL.
^
Z
OMOPHE
cc
CO
f

A

Ju
J,
t









s 	
1 1
20 22 2
                     RETENTION TIME-MINUTES

Figure 1.  Gas chromatogram of haloethers

-------
  COLUMN: TEN AX GC

  PROGRAM:  150°C.-4 MINUTES 16VMINUTE TO 310°C.

  DETECTOR: HALL ELECTROLYTIC CONDUCTIVITY
                      ce
                      LU
                      X
04s        12        16       20


                   RETENTION TIME-MINUTES


Figure 2. Gas chromatogram of haloethers
24

-------
                           CHLORINATED HYDROCARBONS
                                 'METHOD 612

1.  Scope and Application
    1.1   This method covers the determination of certain chlorinated hydro-
         carbons.   The following parameters may be determined by this method:
         Parameter                                             STORET No.
         Hexachlorocyclopentadiene                              34386
         Hexachlorobenzene                                      39700
         Hexachlorobutadiene                                    34391
         Hexachloroethane                                       34396
         1,2-Dichlorobenzene                                    34536
         1,2,4-Drichlorobenzene                                 34551
         1,3-Dichlorobenzene                                    34566
         1,4-Dichlorobenzene                                    34571
         2-chloronaphthalene                                    34581
    1.2   This method is applicable to the determination of these compounds
         in municipal and industrial  discharges.   It is designed to be used
         to meet the monitoring requirements of the National  Pollutant
         Discharge Elimination System (NPDES).   As such, it presupposes a
         high expectation of finding  the specific compounds of interest.  If
         the user  is attempting to screen samples for any or all of the
         compounds above, he must develop independent protocols for the
         verification of identity.
    1.3   The sensitivity of this method is usually dependent upon the level
         of interferences rather than instrumental limitations.  The limits

-------
         of detection listed in Table I represent sensitivities that can be
         achieved in wastewaters in the absence of interferences.
    1.4  This method is recommended for use only by experienced residue
         analysts or under the close supervision of such qualified persons.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
            ?
         using separatory funnel techniques.  The extract is dried by
         passing through a sodium siilfate column and concentrated to a
         volume of 10 ml or less.  Chromatographic conditions are described
         which allow for the accurate measurement of the compounds in the
         extract.
    2.2  If interferences are encountered or expected, the method provides a
         selected general purpose cleanup procedure to aid the analyst in
         their elimination.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
         in all-glass systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the :^.=-s-:ty of the
         industrial complex or municipality being si""lea.  While general
         clean-up techniques are provided as part OT" :--":. method, unique
                                •jr-

-------
         samples may require additional cleanup approaches to achieve the
         sensitivities stated in Table 1.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1  Grab sample bottle - amber glass, 1-lfter or 1-quart
                volume.  French or Boston Round design is recommertded.  The
                container must be washed and solvent rinsed before use "to
                minimize interferences.
         4.1.2  Bottle caps - Threaded to screw on to the sample bottles.
                Caps must be lined with Teflon.  Foil may be substituted if
                sample is not corrosive and the foil is found to be inter-
                ference free.
         4.1.3  Compositing equipment - Automatic or manual compositing
                system. Must incorporate glass sample containers for the
                collection of a minimum of 250 ml.  Sample containers must
                be kept refrigerated during sampling.  No tygon or rubber
                tubing may be used in the system.
    4.2  Separatory funnel - 2000 ml, with Teflon stopcock.
    4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
         frit.
    4.4  Kuderna-Danish (K-D) Apparatus
         4.4.1  Concentrator tube - 10 ml, graduated (Kontes K-570050-1025
                or equivalent).  Calibration must be checked.  Ground glass
                stopper (size 19/22 joint) is used to prevent evaporation of
                extracts.
         4.4.2  Evaporative flask - 500 ml (Kontes K-57001-0500 or
                               nt

-------
                equivalent).   Attach to concentrator tube with springs.
                (Kontes K-662750-0012).
         4.4.3  Snyder column - three-ball macro (Kontes K503000-0121 or
                equivalent).
         4.4.4  Snyder column - two-ball micro (Kontes K-569001-0219 or
                equivalent).
         4.4.5  Boiling chips - solvent extracted,  approximately 10/40 mesh.
    4.5  Water bath - Heated, with Concentric ring  cover, capable of
         temperature control  (_+ 2°C).  The bath should be used in a hood.
    4.6  Gas chromatograph -  Analytical system complete with gas chromato-
         graph suitable for on-column injection and all required accessories
         including electron capture detector, column supplies, recorder,
         gases, syringes.  A data system for measuring peak areas is recom-
         mended.
    4.7  Chromatography column - 300 mm long x 10 mm ID with coarse fritted
         disc at bottom and Teflon stopcock.
5.  Reagents
    5.1  Preservatives:
         5.1.1  Sodium hydroxide - (ACS) 10 N in distilled water.
         5.1.2  Sulfuric acid - (ACS) Mix equal volumes of cone. HgSO^
                with distilled water.
    5.2  Methylene chloride,  Hexane and Petroleum ether (boiling range
         30-60°C) - Pesticide quality or equivalent.
    5.3  Sodium sulfate - (ACS) Granular, anhydrous (purified by heating at
         400°C for 4 hrs. in a shallow tray).
    5.4  Stock standards - Prepare stock standard solutions at a concentra-

-------
         tion of 1.00 ug/ul by dissolving 0.100 grams of assayed reference
         material in pesticide quality isooctane or other appropriate
         solvent and diluting to volume in a 100 ml ground glass stoppered
         volumetric flask.  The stock solution is transferred to ground
         glass stoppered reagent bottles, stored in a refrigerator, and
         checked frequently for signs of degradation or evaporation,
         especially just prior to preparing working standards from them.
    5.5  Florisil - PR grade (60/100 mesh); purchase activated at 1250°F
         and store in the dark in glass containers with glass stoppers or
         foil-lined screw caps.  Before use, activate each batch at 130°C
         in foil-covered glass containers.
6.  Calibration
    6.1  Prepare calibration standards that contain the compounds of
         interest, either singly or mixed together.  The standards should be
         prepared at concentrations covering two or more orders of magnitude
         that will completely bracket the working range of the chromato-
         graphic system.  If the sensitivity of the detection system can be
         calculated from Table I as 100 ug/1 in the final extract, for
         example, prepare standards at 10 ug/1, 50 ug/1, 100 ug/1, 500 ug/1,
         etc. so that injections of 1-5 ul of each calibration standard will
         define the linearity of the detector in the working range.
    6.2  Assemble the necessary gas chromatographic apparatus and establish
         operating parameters equivalent to those indicated in Table I.  By
         injecting calibration standards, establish the sensitivity limit of
         the detector and the linear range of the analytical system for each
         compound.
                              Tf-

-------
    §.3  The cleanup procedure in Section 10 utilizes Florisil chromatog-
         raphy.  Florisil  from different batches or sources may vary in
         absorption capacity.   To standardize the amount of Florisil which
         is used, the use  of lauric acid value (Mills, 1968) is suggested.
         The referenced procedure determines the adsorption from hexane
         solution of lauric acid (mg)  per. gram Florisil.  The amount of
         Florisil to be used for each  column is calculated by dividing this
         ratio by 110 and  multiplying  by 20 grams.
    6.4  Before using any  cleanup procedure, the analyst must process a
         series of calibration standards through the procedure to validate
         elution patterns  and  the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any samples, the analyst should demonstrate
        /
         through the analysis  of a distilled water  method blank, that all
         glassware and reagents are interference-free.  Each time a set of
         samples is extracted  or there is a change  in reagents, a method
         blank should be processed as  a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality  assurance practices should be used with this
         method.  Field replicates should be collected to validate the
         precision of the  sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to validate the accuracy of the analysis.
         Where doubt exists over the identification of a peak on the
         chromatogram, confirmatory techniques such as mass spectroscopy
         should be used.
                              m-

-------
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers, leaving a
         minimum headspace.  Conventional sampling practices should be
         followed, except that the bottle must not be prewashed with sample
         before collection.  Composite samples should be collected in
         refrigerated glass containers in accordance with the requirements
         of the program.  Automatic sampling equipment must be free of tygon
         and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of col-
         lection until extraction.  Chemical preservatives should not be
         used in the field unless more than 24 hours will elapse before
         delivery to the laboratory.  If the samples will not be extracted
         within 48 hours of collection, the sample should be adjusted to a
         pH range of 6.0-8.0 with sodium hydroxide or sulfuric acid.
    8.3  All samples should be extracted immediately and must be extracted
         within 3 days and completely analyzed within 30 days of collection.
9.  Sample Extraction
    9.1  Mark the water meniscus on the side of the sample bottle for later
         determination of sample volume.  Pour the entire sample into a
         two-liter separatory funnel.  Check the pH of the sample with
         wide-range pH paper and adjust to within the range of 5-9 with
         sodium hydroxide or sulfuric acid.
    9.2  Add 60 ml methylene chloride to the sample bottle, seal, and shake
         30 seconds to rinse the inner walls.  Transfer the solvent into the
         separatory funnel, and extract the sample by shaking the funnel for
         two minutes with periodic venting to release vapor pressure.  Allow

-------
     the organic layer to separate from the water phase for a minimum of
     ten minutes.   If the emulsion interface between layers is more than
     one-third the size of the solvent layer, the analyst must employ
     mechanical techniques to complete the phase separation.   The opti-
     mum technique depends upon the sample, but may include stirring,
     filtration of the emulsion through glass wool,-or centrifugation.
     Collect the methylene chloride extract in a 250-ml Erlenmeyer flask.
9.3  Add a second 60-ml volume of methylene chloride to the sample
     bottle and complete the extraction procedure a second time,
     combining the extracts in the Erlenmeyer flask.
9.4  Perform a third extraction in the same manner.  Pour the combined
     extract through a drying column containing 3-4 inches of anhydrous
     sodium sulfate, and collect it in a 500-ml Kuderna-Danish (K-D)
     flask equipped with a 10 ml concentrator tube.  Rinse the
     Erlenmeyer flask and column with 20-30 ml methylene chloride to
     complete the quantitative transfer.
9.5  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus  on a hot
     water bath (60-65°C) so that the concentrator tube is partially
     immersed in the hot water, and the entire lower rounded  surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distillation
     the balls of the column will actively chatter but the chambers will
     not flood.  When the apparent volume of liquid reaches 1-2 ml,

                             JPI17

-------
         remove the K-D apparatus and allow it to drain for at least 10,-
         minutes while cooling.  NOTE:  The dichlorobenzenes have .a,suf-
         ficiently high volatility that signficant losses may occur in
         concentration steps if care is not exercised.  It is important to
         maintain a constant gentle evaporation rate and not to allow the
         liquid volume to fall  below 1-2 ml before removing the K-D from the
         hot water bath.
    9.6  Momentarily remove the Snyder column, add 50 ml hexane and a new
         boiling chip and replace the column.   Raise the temperature of the
         water bath to 85-90°C.  Concentrate the extract as in 9.5, except
         using hexane to prewet the column.  Remove the Snyder column and
         rinse the flask and its lower joint into the concentrator tube with
         1-2 ml of hexane.   A 5-ml  syringe is  recommended for this operation.
         Stopper the concentrator tube and store refrigerated if further
         processing will not be performed immediately.
    9.7  Determine the original sample volume by refilling the sample bottle
         to the mark and transferring the liquid to a 1000 ml graduated
         cylinder.  Record  the  sample volume to the nearest 5 ml.
    9.8  Unless the sample  is known to require cleanup, proceed to analysis
         by gas chromatography.
10.  Cleanup and Separation
    10.1  Florisil Column Cleanup for Chlorinated Hydrocarbons.
         10.1.1 Adjust the  sample extract to 10 ml.
         10.1.2 Place a 12  gram charge of activated Florisil  (see 6.3)  in a
                10 mm ID chromatography column.   After settling the Florisil
                by tapping  the  column, add a 1-2 cm layer of anhydrous

-------
                granular sodium sulfate to the top.
         10.1.3  Pre-elute the  column,  after cooling,  with 100 ml  of
                petroleum ether.   Discard the eluate  and just prior to
                exposure of the sulfate layer to air, quantitatively
                transfer the sample extract into the  column by decantation
                and subsequent petroleum ether washings.  Discard the
                eluate.   Just  prior to exposure of the sodium sulfate layer
                to the air, begin eluting the column  with 200 ml  petroleum
                ether and collect the  eluate in a 500 ml K-D flask equipped
                with a 10 ml concentrator tube.  This fraction should
                contain  all of the chlorinated hydrocarbons.
         10.1.4  Concentrate the fraction by K-D as in 9.5 except  prewet the
                column with hexane. When the apparatus is cool,  remove the
                Snyder column  and rinse the flask and its lower joint into
                the concentrator tube  with 1-2 ml hexane.  Analyze by gas
                chromatography.
11.  Gas Chromatography
    11.1 Table I summarizes the recommended gas chromatographic column
         materials and operating conditions for the instrument.  Included in
         this table are  estimated retention times and sensitivities that
         should  be achieved by this method.  Examples of the separations
         achieved by this column are shown in Figures 1 and 2.  Calibrate
         the system daily with a minimum of three injections of calibration
         standards.
    11.2 Inject  2-5 ul of the sample extract using the solvent-flush tech-
         nique.   Smaller (1.0 ul) volumes can be injected if automatic
                                  jr-

-------
         devices are employed.  Record the volume injected to the nearest
         0.05 ul, and the resulting peak size, in area units.
    11.3 If the peak area exceeds the linear range of the system, dilute the
         extract and reanalyze.
    11.4 If the peak area measurement is prevented by the presence of inter-
         ferences, further cleanup is required.
12.  Calculations
    12.1 Determine the concentration of individual compounds according to
         the formula:
         Concentration, ug/1 = (A) (B) (Vt)
                                 (Vi)  (Vs)
         where A = Calibration factor for chromatographic system, in
                   nanograms material per area unit.
               B = Peak size in injection of sample extract, in area units
               V.. = volume of extract injected (ul)
               V. = Volume of total extract (ul)
               V  = Volume of water extracted (ml)
    12.2 Report results in micrograms per liter without correction for
         recovery data.  When duplicate and spiked samples are analyzed, all
         data obtained should be reported.
13.  Accuracy and Precision
    The U.S. EPA Environmental Monitoring and Support Laboratory in
    Cincinnati is in the process of conducting an interlaboratory method
    study to determine the accuracy and precision of this test procedure.

-------
                                 BIBLIOGRAPHY
1.  "Development and Application of Test Procedures for Specific Organic
    Toxic Substances in Wastewaters.  Category 3 - Chlorinated Hydrocarbons
    and Category 8 - Phenols."  Report for EPA Contract 68-03-2625 (In
    preparation).
2.  Mills, P.A., "Variation of Florisil Activity:  Simple Method for
    Measuring Absorbent Capacity and Its Use in Standardizing Florisil
    Columns," Journal of the Association of Official Analytical Chemists,
    51, 29 (1968).

-------
                                   TABLE I

                Gas Chromatography of Chlorinated Hydrocarbons
Compound
Retention Time (min.)

Column 1
Detection

Limit (ug/1)
1 ,3-dichlorobenzene
1 , 4-d i ch 1 orobenzene
hexachloroethane
1,2-dichlorobenzene
hexachlorobutadiene
1,2,4-trichlorobenzene
hexachlorocyclopentadiene
2-chloronaphthalene
hexachlorobenzene
4.0
4.3
4.8
5.3
11.6
12.4
1.5*
2.5*
7.0*
0.009
0.018
0.001
0.012
0.001
0.006
0.001
0.015
0.001
Column 1 conditions:  Gas Chrom Q 80/100 mesh coated with 1.5% OV-1/1.5%
    OV-225 packed in a 1.8 m long x 2 mm ID glass column with 5% Methane/95%
    Argon carrier gas at 30 ml/min flow rate.  Column temperature is 75°C
    except where * indicates 160°C.  Under these conditions R.T. of Aldrin
    is 18.8 minutes at 160°C.

 Detection limit is calculated from the minimum detectable GC response of
    the electron capture detector being equal to five times the GC back-
    ground noise, assuming a 10 ml final volume of the 1 liter sample
    extract, and assuming a GC injection of 5 micro!iters.

-------
           COLUMN:  1.5% OV-U 1.5% OV-225 ON GAS CHROM Q
           TEMPERATURE:  75°C.
           DETECTOR:  ELECTRON CAPTURE
                                   (VI

                                   LU
                                   QQ
                                   O
                                   cc
                                   O
                                   _i
                                   X
                                   O
                                   {VI
                 8       12       16
                 RETENTION TIME-MINUTES
20
Figure 1. Gas chromatogram of chlorinated hydrocarbons
                 T-

-------
      o
      <
      Q.

      O
      o
      cc
      O

      z
      o

      x
      LU
      z
COLUMN:  1.5°i OV-1

         1.5% OV-225 ON GAS CHROM Q

TEMPERATURE: 160°C.

DETECTOR: ELECTRON CAPTURE
      LU
                         CO

                         O
                         cc

                         O


                         I

                         O
                  8       12      16


             RETENTION TIME-MINUTES
Figure 2.  Gas chromatogram cf :;h'orinated hydrocarbons

-------
                     2,3,7,8-TETRACHLORODIBENZO-p-DIOXIN
                                  METHOD 613

1.   Scope and Application
    1.1   This method covers the determination of 2,3,7,8-tetrachlorodibenzo-
         p-dioxin (TCDD).  The following parameter may be determined by this
         method:
                Parameter                            STORET No.
                  TCDO                                 34675

    1.2' This method is applicable to the determination of TCDD in municipal
         and industrial discharges.  It is designed to be used to meet the
         monitoring requirements of the National Pollutant Discharge
         Elimination System (NPDES).  As such, it presupposes the potential
         for finding trace levels of TCDD in the sample.  The method incor-
         porates techniques that can also be used to screen samples for TCDD
         using an electron capture detector.
    1.3  The sensitivity of this method is usually dependent upon the level
         of interferences rather than instrumental limitations.  The limit
         of detection listed in Table I represents the sensitivity that can
         be achieved in wastewaters in the absence of interferences.
    1.4  This method is recommended for use only by analysts experienced
         with residue analysis and skilled in mass spectral analytical
         techniques.
    1.5  Because of the extreme toxicity of this compound, the analyst must
         take elaborate precautions to prevent exposure to himself, or to
                              3-

-------
         others, of materials known or believed to contain TCDD.  The
         appendix to this method contains guidelines and protocols that
         should serve as minimum safe-handling standards for the laboratory.
2.  Summary of Method
    2.1  A 1-liter sample of wastewater is extracted with methylene chloride
         using separatory funnel techniques.  The extract is dried and ex-
         changed to hexane while being concentrated to a volume of 1.0 ml or
         lower.  Capillary column GC/MS conditions and internal standard
         techniques are described which allow for the measurement of TCDD in
         the extract.  Electron capture gas chromatographic conditions are
         also provided to permit the analyst to use this equipment to pre-
         screen samples before GC/MS analysis.
    2.2  If interferences are encountered, the method provides selected
         general purpose cleanup procedures to aid the analyst in their
         elimination.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the
         conditions of the analysis by running method blanks.  Specific
         selection of reagents and purification of solvents by distillation
         in all-glass systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the
         industrial complex or municipality being sampled.  TCDD is often

-------
         associated with other interfering chlorinated compounds which are
         at concentrations several magnitudes higher than that of TCDD,
         While general cleanup techniques are provided as part of this
         method,  unique samples may require additional cleanup approaches to
         achieve the sensitivity stated in Table 1.
    3.3  The other isomers of tetrachlorodibenzo-p-dioxin may interfere with
         the measurement of 2,3,7,8-TCDD.  Capillary column gas chromatog-
         raphy is required to resolve those isomers that yield virtually
         identical mass fragmentation patterns.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling. •
         4.1.1  Grab sample bottle - amber glass, 1-liter or 1-quart
                volume.  French or Boston Round design is recommended.  The
                container must be washed and solvent rinsed before use to
                minimize interferences.
         4.1.2  Bottle caps - Threaded to screw on to the sample bottles.
                Caps must be lined with Teflon.  Foil may be substituted if
                sample is not corrosive.
         4.1.3  Compositing equipment - Automatic or manual compositing
                system. Must incorporate glass sample containers for the
                collection of a minimum of 250 ml.  Sample containers must
                be kept refrigerated during sampling.  No tygon or rubber
                tubing may be used in the system.
    4.2  Separatory funnels - 2000 ml and 500 ml, with Teflon stopcock.
    4.3  Drying column - 20 mm ID pyrex chromatographic column with coarse
         frit.

-------
4.4  Kuderna-Danish (K-D) Apparatus
     4.-4.1  Concentrator tube - 10 ml, graduated (Kontes K-570050- 1025
            or equivalent).  Calibration must be checked.  Ground glass
            stopper (size 19/22 joint) is used to prevent evaporation of
            extracts.
     4.4.2  Evaporative flask - 500 ml (Kontes K-57001-0500 or equiva-
            lent).  Attach to concentrator tube with springs.  (Kontes
            K-662750-0012).
     4.4.3  Snyder column - three-ball macro (Kontes K503000-0121 or
            equivalent).
     4.4.4  Snyder column - two-ball micro (Kontes K-569001-0219 or
            equivalent).
     4.4.5  Boiling chips - solvent extracted, approximately 10/40 mesh.
4.5  Water bath - Heated, with concentric ring cover, capable of
     temperature control (+ 2°C).  The bath should be used in a hood.
4.6  Gas chromatograph - Analytical system complete with gas chromato-
     graph suitable for on-column injection and all required accessories
     including electron capture, packed and capillary column supplies,
     recorder, gases,  syringes.  A data system for measuring peak areas
     is recommended.
4.7  GC/Mass Spectrometer system - electron ionization source, capable
     of selected ion monitoring in groups of two or more ions, and
     related data system.
4.8  Chromatography column - 300 mm long x 10 mm ID with coarse fritted
     disc at bottom and Teflon stopcock.
4.9  Chromatography column - -CO mm long x 11 mm ID with coarse fritted
                                 ?£>

-------
         disc at bottom and Teflon stopcock.
    4.10 Pipets - Disposable, Pasteur, 150 mm long x 5 mm ID (Fisher
         Scientific Co., No. 13-678-6A or equivalent).
5.  Reagents
    5.1  Sodium hydroxide - (ACS) 10 N  and 1 N in distilled water.  Wash
         the solutions with methylene chloride and with hexane.
    5.2  Sulfuric acid - (ACS) (HI) and IN.  For (1+1), mix equal volumes
         of cone. HgSO^ with distilled water.  Wash the solutions with
         methylene chloride and with hexane.
    5.3  Methylene chloride, hexane, benzene, tetradecane - Pesticide
         quality or equivalent.
    5.4  Sodium Sulfate - (ACS) Granular, anhydrous (purified by heating at
         400°C for 4 hrs. in a shallow tray).
    5.5  Stock standards - In a glovebox, prepare stock standard solutions
         of TCDD and    Cl-TCDD.  The stock solutions are stored in the
         glovebox, and checked frequently for signs of degradation or evapo-
         ration, especially just prior to preparing working standards from
         them.
    5.6  Silica gel - high purity grade, 100/120 mesh, (Fisher Scientific
         Co., No. S-679 or equivalent).
    5.7  Alumina - neutral, 80/200 mesh (Fisher Scientific Co., No. A-540 or
         equivalent).  Before use activate for 24 hours at 130°C in a foil-
         covered glass container.
    5.8  Activated Coconut Charcoal - 50/200 mesh (Fisher Scientific Co.,
         No. 5-690A or equivalent).
6.  Calibration

-------
    6.1  Prepare GC/MS calibration standards for the internal standard
         technique that will  allow for measurement of relative response
         factors of at least  three TCDD/37C1 TCDD ratios.  The 37C1-TCDD
         concentration in the standards should be fixed and selected to
         yield a reproducible response at the most sensitive setting of the
         mass spectrometer.
    6.2  Assemble the necessary GC or GC/MS apparatus and establish
         operating parameters equivalent to those indicated in Table I.
         Calibrate the GC/MS  system according to Eichelberger, et al.
         (1975).  By injecting calibration standards, establish the response
         factors for TCDD vs. 37C1-TCDD.
    6.3  Before using any cleanup procedure, the analyst must process a
         series of calibration standards through the procedure to validate
         elution patterns and the absence of interferences from the reagents.
7.  Quality Control
    7.1  Before processing any samples, the analyst should demonstrate
         through the analysis of a distilled water method blank, that all
         glassware and reagents are interference-free.   Each time a set of
         samples is extracted or there is a change in reagents, a method
         blank should be processed as a safeguard against chronic laboratory
         contamination.
    7.2  Standard quality assurance practices should be used with this
         method.  Field replicates should be collected  to validate the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to validate the precision of the analysis.  Fortified
         samples should be analyzed to validate the accuracy of the analy-

-------
         sis, although surrogate spikes are recommended because of the
         toxicity of TCDD.  Where doubt exists over the identification of a
         peak on the electron capture chromatogram, mass spectroscopy must
         be used for clarification or confirmation.
8.  Sample Collection, Preservation, and Handling
    8.1  Grab samples must be collected in glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prewashed with sample before collection.  Composite samples
         should be collected in refrigerated glass containers in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of col-
         lection until extraction.  Chemical preservatives should not be
         used in the field unless more than 24 hours will elapse before
         delivery to the laboratory.  If the samples will not be extracted
         within 48 hours of collection, the sample should be adjusted to a
         pH range of 6.0-8.0 with sodium hydroxide or" sulfuric acid.
    8.3  All samples must be extracted within 7 days and completely analyzed
         within 30 days of collection.
9.  Sample Extraction
    CAUTION:  If there is a remote possibility that the sample contains TCDD
    at measurable levels, all of the following operations must be performed
    in a limited access laboratory with the analyst wearing full protective
    covering for all exposed skin surfaces.  See Appendix.
    9.1  Mark the water meniscus on the side of the sample bottle for later
         determination of sample volume.  Pour the entire sample into a
                            HE-

-------
     two-liter separatory funnel.  Check the pH of the  sample with
     wide-range pH paper and adjust to within the range of 5-9 with
     sodium hydroxide or sulfuric acid.
9.2  Add 60 ml methylene chloride to the sample bottle, seal, and shake
     30 seconds to rinse the inner walls.  Transfer the solvent  into the
     separatory funnel, and extract the sample by shaking the funnel for
     two minutes with periodic venting to release vapor pressure.  Allow
     the organic layer to separate from the water phase for a minimum of
     ten minutes.  If the emulsion interface between layers is more than
     one-third the size of the solvent layer, the analyst must employ
     mechanical techniques to complete the phase separation.  The
     optimum technique depends upon the sample, but may include  stir-
     ring, filtration of the emulsion through glass wool, or centrifu-
     gation.  Collect the methylene chloride extract in a 500-ml
     separatory funnel.
9.3  Add a second 60-ml volume of methylene chloride to the sample
     bottle and complete the extraction procedure a second time,
     combining the extracts in the 500-ml separatory funnel.
9.4  Perform a third extraction in the same manner.  To the combined
     extracts in the separatory funnel add 1CJD ml IN NaOH.  Shake the
     funnel for 30-60 seconds.  Allow the layers to separate and draw
     the organic layer into a 250 ml  Erlenmeyer flask.   Discard  the
     aqueous layer and return the organic layer to the separatory
     funnel.  Perform a second wash of the organic layer with 1  N NaOH
     and discard the aqueous layer.
9.5  In the same manner wash the organic layer twice with 100 ml IN

-------
     h^SO^, discarding the aqueous layers.
9.6  Wash the organic layer three times with 100 ml hLO, discarding
     the aqueous layers.
9.7  Pour the organic layer extract through a drying column containing
     3-4 inches of anhydrous sodium sulfate, and collect it in a 500-ml
     Kuderna-Danish (K-D) flask equipped with a 10 ml concentrator tube.
     Rinse the Erlenmeyer flask and column with 20-30 ml methylene
     chloride to complete the quantitative transfer.
9.8  Add 1-2 clean boiling chips to the flask and attach a three-ball
     Snyder column.  Prewet the Snyder column by adding about 1 ml
     methylene chloride to the top.  Place the K-D apparatus on a hot
     water bath (60-65°C) so that the concentrator tube is partially
     immersed in the hot water, and the entire lower rounded surface of
     the flask is bathed in vapor.  Adjust the vertical position of the
     apparatus and the water temperature as required to complete the
     concentration in 15-20 minutes.  At the proper rate of distil-
     lation, the balls of the column will actively chatter but the
     chambers will not flood.  When the apparent volume of liquid
     reaches 1 ml, remove the K-D apparatus and allow it to drain for at
     least 10 minutes while cooling.
9.9  Momentarily remove the Snyder column, add 50 ml hexane and a new
     boiling chip  and replace the Snyder column.  Increase the tempera-
     ture of the water bath to 80°C.  Prewet the Snyder column by
     adding about  1 ml hexane to the top.  Evaporate the solvent as in
     9.8.   Remove the Snyder column and rinse the flask and its lower
     joint into the concentrator tube with 1-2 ml of hexane.  A 5-ml
                           at-

-------
     syringe is recommended for this operation.
9.10 Add a clean boiling chip and attach a micro-Snyder column.  Prewet
     the column by adding about 1 ml hexane to the top.  Place the K-D
     apparatus on the 80°C water bath so that the concentrator tube is
     partially immersed in the hot water.  Adjust the vertical position
     of the apparatus and the water temperature as required to complete
     the concentration in 5-10 minutes.  At the proper rate of distil-
     lation the balls of the column will actively chatter but the
     chambers will not flood.  When the apparent volume of liquid
     reaches about 0.5 ml, remove the K-D apparatus and allow it to
     drain for at least 10 minutes while cooling.  Remove the micro
     Snyder column and rinse its lower joint into the concentrator tube
     with 0.2 ml hexane.  Adjust the extract volume to 1.0 ml with
     hexane.  Stopper the concentrator tube and store refrigerated if
     further processing will not be performed immediately.
9.11 Determine the original sample volume by refilling the sample bottle
     to the mark and transferring the liquid to a 1000 ml graduated
     cylinder.  Record the sample volume to the nearest 5 ml.
9.12 The analyst has several options available to him depending upon the
     nature of the sample and the availability of resources:
     9.12.1 If the appearance of the extract or previous experience with
            the matrix indicates cleanup will be required the analyst
            should proceed with one or more techniques as described in
            paragraph 10.
     9.12.2 If the analyst wishes to screen the sample for the possible
            presence of TCDD before GC/MS analysis, he can analyze the

-------
                extract by packed column or capillary column electron
                capture detection, as in paragraph 11.
         9.12.3 The analyst may proceed directly to GC/MS analysis, para-
                graph 12.
10.  Cleanup and Separation
    10.1 Several cleanup column choices are offered to the analyst in this
         section.  Although any of them may be used independently, the
         silica gel column (10.2) followed immediately by the alumina column
         .(10.3) has been used frequently to overcome background problems
         encountered by the GC/MS.
    10.2 Silica Gel Column Cleanup for TCDD.
         10.2.1.Fill a 400 mm long x 11 mm ID chromatography column with
                silica gel to the 300 mm level, tapping the column gently to
                settle the silica gel.  Add 10 mm anhydrous sodium sulfate
                to the top of the silica gel.
         10.2.2 Preelute the column with 50 ml 20% benzene/80% hexane (V/V).
                Adjust the elution rate to 1  ml/min.  Discard the eluate and
                just prior to exposure of the sodium sulfate layer to the
                air, transfer the entire 1.0 ml sample extract onto the
                column, using two 2 ml portions of 20% benzene/80% hexane to
                complete the transfer.
         10.2.3 Just prior to exposure of the sodium sulfate layer to the
                air, add 40 ml 20% benzene/80% hexane to the column.
                Collect the eluate in a 500 ml K-D flask equipped with a 10
                ml concentrator tube.
         10.2.4 Evaporate the fraction to 1.0 ml by standard K-D techniques
                                 It

-------
            (9.8-9.10).  Analyze by ECGC (11), GC/MS (12) or continue
            cleanup as described below (10.3).
10.3 Alumina Column Cleanup for TCDD
     10.3.1  If the extract is not in hexane, add 0.1-0.2 ml tetradecane
            keeper and concentrate it at room temperature down to this
            volume using a stream of dry nitrogen gas.   Dilute'to 1.0 ml
            with hexane.
     10.3.2  Fill a 300 mm long x 10 mm ID chromatography column with
            activated alumina to the 150 mm level, tapping the column
            gently to settle the alumina.  Add 10 mm anhydrous sodium
            sulfate to the top of the alumina.
     10.3.3  Preelute the column with 50 ml  hexane.  Adjust the elution
            rate to 1 ml/min.  Discard the  eluate and just prior to
            exposure of the sodium sulfate  layer to the air, transfer
            the entire 1 ml sample extract  onto the column, using two
            additional 2 ml portions of hexane to complete the transfer.
     10.3.4  Just prior to exposure of the sodium sulfate layer to the
            air, add 50 ml 3% methylene chloride/97% hexane (V/V) and
            continue the elution of the column.  Discard the eluate.
     10.3.4  Next elute the column with 50 ml 20% methylene chloride/80%
            hexane (V/V) into a 500 ml K-D  flask equipped with a 10 ml
            concentrator tube.  Concentrate the collected fraction to
            1.0 ml by standard K-D technique (9.8-9.10).  Analyze by
            ECGC (11), GC/MS (12) or continue cleanup as described below
            (10.4).
10.4 Charcoal and Silica-gel Column Cleanup for TCDD.

-------
         10.4.1  Prepare a homogeneous mixture of 1 part activated charcoal
                to 140 parts silica-gel.   Fill a 5 mm ID disposable pipet to
                a length of 50 mm,  tapping the column to settle the mixture.
         10.4.2  Preelute the column with  5 ml hexane.  Discard the eluate
                and just prior to exposure of the top of the column to the
                air, transfer an 0.5 ml  aliquot of sample extract onto the
                column, using an additional 0.5 ml hexane to complete the
                transfer.
         10.4.3  Just prior to exposure of the top of the column to the air,
                add 10 ml hexane and continue the elution of the column.
                Discard the eluate.
         10.4.4  Next,  elute the column with 10 ml benzene into a 10 ml K-D
                concentrator tube.   Concentrate the eluate to 1.0 ml with
                micro-K-D concentration  (9.10) on a boiling water bath.
                Analyze by ECGC (11) or GC/MS (12).
11.  Electron Capture Screening
    11.1  The sample extracts can be screened by electron capture gas chroma-
         tography at the option of the analyst in an effort to reduce the
         workload on the GC/MS system.  Either packed or capillary column
         techniques may be used for this  purpose.  The only acceptable
         conclusions that can be reached  with this technique are:  (a) TCDO
         is not  detectable at the detection limit of the procedure:  (b)
         TCDD is not present above a stated concentration or control level:
         and (c) the presence or absence  of TCDD is unresolved.
    11.2  Table I summarizes some recommended gas chromatographic column
         materials and operating conditions for the instrument.  Included in
                                 IT- IS 5"

-------
         this table are estimated retention times and sensitivities that
         should be achieved by this method.  An example of the chromatog-
         raphy achieved by the packed column is shown in Figure 1 and by the
         capillary column in Figure 2.  Calibrate the system daily with a
         minimum of three injections of calibration standards.
    11.3 For packed column GC, inject 2-5 ul of the sample extract using the
         solvent-flush technique.  A splitless injector is recommended for
         the capillary system, but for optimum peak geometry a solvent
         exchange to tetradecane is required (10.3.1).  Record the volume
         injected to the nearest 0.05 ul, and the resulting peak size, in
         area units.
    11.4 If there is no measureable baseline deflection at the retention
         time of TCDD, report the result as less than the detection limit of
         the electron capture system.
    11.5 If a measurable peak appears within the tolerances of the TCDD
         retention time of the system, the analyst should proceed to GC/MS
         (12).
    11.6 If the complexity of the chromatogram defies interpretation, the
         analyst may want to pursue cleanup (10) followed by reanalysis by
         ECGC, or proceed directly to GC/MS.
12.  GC/MS Analysis
    12.1 Table I summarizes the recommended capillary column gas chromato-
         graphic materials and operating conditions for the instrument.
         Included in this table is the estimated retention time and
         sensitivity that should be achieved by this method.  An example of
         the chromatography achieved by this column is shown in Figure 2.

-------
     Calibrate the system dally, with a minimum at three Injections of
     standard mixtures.
12.2 Add a known amount of   Cl-TCOO to the sample extract.
12.3 Analyze samples with selected ion monitoring of at least two Ions
     characteristic of TCDD (m/e 320 and m/e 322) and of   CI-TCOD
     (m/e 328).  Proof of the presence of TOO exists if the following
     conditions are met:
     12.3.1  The retention time of the peak 1n the sample must match
             that in the standard, within the performance specifications
             of the analytical system.
     12.3.2  The ratio of ions (320:322) must agree within 1W of that
             in the standard.
     12.3.3  The retention time of the peak maximum for the m/e 320 peak
             must exactly match that of the 322 peak.
12.4 Quantitate the TCDO peak from the response relative to the.   Cl
     TCDD internal standard.
12.5 If a response 1s obtained for both ions but is outside the expected
     ratio, then a co-eluting impurity may be suspected.  In this case,
     another set of ions characteristic of the TCDD molecule, should  be
     analyzed.  A good choice of ions 1s m/e 257 and m/e 259.  These
     ions are useful in characterizing the molecular structure of TCDO.
     Suspected impurities such as DDE, ODD or PCS residues can be
     confirmed by checking for their major fragments.  These materials
     can be removed by the cleanup columns.  If available, an analysis
     of the EC chromatogram will provide  insight into  the complexity  of
     the problem and will determine the manner in which the mass
     spectrometer will be used.

-------
    12.6 If broad background interference restricts the sensitivity of the
         GC/MS analysis, the analyst should employ cleanup procedures (10)
         and reanalyze by GC/MS.
    12.7 In those circumstances where these procedures do not yield a
         definitive conclusion, then the use of high resolution mass
         spectrometry is suggested.
13. Calculations
    13.1 Determine the concentration of individual compounds according to
         the formula: •

         Concentration,  ug/1 =     (A) (V.)
                                  (V )1 (Vs)
         where A = Nanograms TCDD injected into the GC/MS from the
                   calibration curve.
               V. = volume of extract injected (ul)
               Vt = Volume of total  extract (ul)
               V  = Volume of water  extracted (ml)
    13.2 Report results in micrograms per liter without correction for
         recovery data.  When duplicate and spiked samples are analyzed, all
         data obtained should be reported.
14.  Accuracy and Precision
    No data available at this time.
                                   inn-

-------
                                 BIBLIOGRAPHY

1.   "Development and Application of Test Procedures for Specific Organic
    Toxic Substances in Wastewaters.  Category 6-Dioxin."  Report for EPA
    Contract 68-03-2635 (In preparation).

2.   "Reference Compound to Calibrate Ion Abundance Measurements in Gas
    Chromatography--Mass Spectrometry Systems," J. W. Eichelberger, L. E.
    Harris, and W. L. Budde, Anal. Chem. 47, 995-1000 (1975).
                                   '- )

-------
                                   Table I

                          Gas Chromatography of TCDD

                                                                Detection
Column                      Retention Time (min.)               Limit (ug/L)

Packed1                          7.3                             0.003
Glass Capillary 2                9.5                             0.003


^Packed column conditions:  Supelcoport 100/120 mesh coated with 1.5%
    SP-2250/1.95% SP-2401 packed in a 180 cm long x 2 mm ID glass column
    with 5% Methane/95% Argon carrier gas at 25 ml/min flow rate.  Column
    temperature is 220°C.
       capillary column conditions:  SP-2250 coated on a 30 m long x 0.25
    mm ID glass column (Supelco No. 2-3714 or equivalent) with helium
    carrier gas at 30 cm/sec linear velocity run splitless.  Column
    temperature is 210°C.

 Detection limit is calculated from the minimum detectable GC response being
    equal to five times the GC background noise, assuming a 1 ml effective
    final volume of the 1 liter sample extract, and assuming a GC injection
    of 5 micro! iters.  Detection levels apply to both electron capture and
    GC/MS detection.

-------
                                  APPENDIX A
                       Safe Handling Practices for TCDD

    Dow Chemical U.S.A. has issued the following precautions for safe
handling TCDD in the laboratory.  In addition to these practices, the
following points are also helpful:

1.  Contamination of the laboratory will be minimized by conducting all
    manipulations in the hood.

2.  Effluent of the gas chromatography (from the Nickel-63 detector or as a
    result of splitting when capillary columns are used) should pass through
    either a column of activated charcoal or bubbled through a trap contain-
    ing oil or high-boiling alcohols.

3.  Liquid waste can be dissolved in methanol or ethanol and irradiated with
    ultraviolet light with wavelength greater than 290 nm for several days.

                                   * * * *
                       PRECAUTIONS FOR SAFE HANDLING OF
                2,3,7,8-TETRACHLORODIBENZO-PARA-DIOXIN (TCDD)
                      IN THE LABORATORY (REVISED 11/78)
    The following statements on safe handling are as complete as possible  on
the basis of available toxicological information.  The precautions for  safe

-------
handling and use are necessarily general in nature since detailed,  specific
recommendations can be made only for the particular exposure and circum-
stances of each individual use.  Inquiries about specific operations or uses
may be addressed to the Dow Chemical Company.  Assistance in evaluating the
health hazards of particular plant conditions may be obtained from  certain
consulting laboratories and from State Departments of Health or of  Labor,
many of which have an industrial health service.
    TCDD is extremely toxic to laboratory animals.  However, it has been
handled for years without injury in analytical and biological laboratories.
Techniques used in handling radioactive and infectious materials are
applicable to TCDD.
Protective Equipment;  Throw-away plastic gloves, apron or lab coat, safety
glasses and lab hood adequate for radioactive work.
Training;  Workers must be trained in the proper method of getting  out of
contaminated gloves and clothing without contacting the exterior surfaces.
Personal Hygiene;  Thorough washing of hands and forearms after each manipu-
lation and before breaks (coffee, lunch, and shift).
Confinement;  Isolated work area, posted with signs, segregated glassware
and tools, plastic-backed absorbent paper on benchtops.
Waste;  Good technique includes minimizing contaminated waste.  Plastic bag
liners should be used in waste cans.  Janitors must be trained in safe hand-
ling of waste (one accidental case of chloracne resulted from handling
laboratory waste in a routine manner).
Disposal of Wastes;  TCDD decomposes above 800°C.  Low-level waste  such as
the absorbent paper, tissues, animal remains and plastic gloves may be
burned in a good incinerator.  Gross quantities (milligrams) should be

-------
packaged securely and disposed through commerical or governmental channels
which are capable of handling high-level radioactive wastes or extremely
toxic wastes.  Liquids should be allowed to evaporate in a good hood  and  in
a disposable container.  Residues may then be handled as above.
Decontamination:  Personal — any mild soap with plenty of scrubbing  action:
Glassware, Tools, and Surfaces ~ Chlorothene* NU Solvent is the least toxic
solvent shown to be effective.  Satisfactory cleaning may be accomplished by
rinsing with Chlorothene, then washing with any detergent and water.  Dish
water may be disposed to the sewer.  It is prudent to minimize solvent
wastes because they may require special disposal through commercial sources
which are expensive.
Laundry;  Clothing known to be contaminated should be disposed with the
precautions described under "Disposal of Wastes."  Lab coats or other
clothing worn in TCDD work may be laundered.  Clothing should be collected
in plastic bags.  Persons who convey the bags and launder the clothing
should be advised of the hazard and trained in proper handling.  The
clothing may be put into a washer without contact if the launderer knows  the
problem.  The washer should be run through a cycle before being used  again
for  other clothing.
Wipe Tests;  A useful method of determining cleanliness of work surfaces  and
tools is to wipe the surface with a piece of filter paper.  Extraction and
analysis by  gas chromatography can achieve a limit of sensitivity of  0.1
microgram per wipe.  The analytical method is available upon request. Less
than 1 microgram TCDD per  sample  indicates acceptable cleanliness;  anything

*Trademark of the Dow Chemical Company

-------
higher warrants further cleaning.  More than 10 micrograms on  a wipe  sample
indicates an acute hazard and requires prompt cleaning before  further use of
the equipment or work space.  It indicates further that unacceptably  sloppy
work habits have been employed in the past.
Inhalation:  Any procedure that may produce airborne contamination, must be
done with good ventilation.  Gross losses to a ventilation system must .not
be allowed.  Handling of the dilute solutions normally used in analytical
and animal work presents no inhalation hazards except in case  of an accident,
Accidents;  Remove contaminated clothing immediately, taking precautions not
to contaminate skin or other articles.  Wash exposed skin vigorously and
repeatedly until medical attention is obtained.
    For clinical advice, contact B. B. Holder, M.D., Medical Director, Dow
Chemical U.S.A., Midland, Michigan, 48640, (telephone 517/636-2108).  For
detailed safe handling precautions for specific procedures, contact L. G.
Silverstein, Industrial Hygiene Laboratory, Dow Chemical U.S.A., Midland,
Michigan, 48640 (telephone 517/636-1688).
                            331-

-------
      COLUMN: 1.5*6 SP-2250* 1.95% SP-2401 ON SUPELCOPORT
      TEMPERATURE:  220°C
      DETECTOR: ELECTRON CAPTURE
 02468
              RETENTION TIME-MINUTES

Figure 1.  Gas chromatogram of TCDD

-------
 73.3

  320

100.0

  322

 50.1

  324


214.1
EXTRACTED ION CURRENT PROFILE
EXTRACTED ION CURRENT PROFILE
                     (M/e - 322)
EXTRACTED ION CURRENT PROFILE
                     (M/e =324)
      TOTAL ION CURRENT PROFILE
            8               9              10
                   RETENTION TIME-MINUTES

    Figure 2.  Total and extracted ion current profiles of TCDD
             "2T-

-------
                                  METHOD 624
                          ORGANICS BY PURGE AND TRAP
1.   Scope and Application
     1.1  This method is designed to determine those "unambiguous priority
          pollutants," associated with the Consent Decree, that are amenable
          to the purge and trap method.  The parameters listed in Table  1
          may be determined by this method.
     1.2  .This method is applicable to the determination of these compounds
          in municipal and industrial discharges.  It is designed to be  used
          to meet the survey requirements of the Effluent Guidelines
          Division (EGD) and the monitoring requirements of the National
          Pollutants Discharge Elimination System (NPDES).
     1.3  The detection limit of this method is usually dependent upon the
          level of interferences rather than instrumental limitations.   The
          limits listed in Table 2 represent sensitivities that can be
          achieved in wastewaters.
     1.4  The GC/MS parts of this method are recommended for use only by
          persons experienced in GC/MS analysis or under the close
          supervision of such qualified persons.
     1.5  The trapping and chromatographic procedures described do not apply
          to the most volatile of the priority pollutants,
          dichlorodifluoromethane.  An alternative three stage trap
          containing charcoal is to be used -:f th's compound is to be
          analyzed.  See EPA Method 601 and Reference 1.  Primary ion for

-------
          quantitative analysis is 101.  The secondary ions  are 85, 87, and
          103.
     1.6  Although this method can be used for qualitative screening for
          acrolein and acrylonitrile, the purging efficiencies are low and
          erratic.  For quantitative analysis of these compounds, use either
          direct aqueous injection (Ref. 4-6) or EPA Method 603, Acrolein
          and 'Acrylonitrile, EMSL, Cincinnati, Ohio.
2.   Summary of Method
     2.1  A sample of wastewater is purged with a stream of  inert gas.  The
          gas is bubbled through a 5 ml water sample contained in a
          specially designed purging chamber.  The volatile organics are
          efficiently transferred from the aqueous phase into the gaseous
          phase where they are passed through a sorbent bed designed to trap
          out the organic volatiles.  After purging is complete, the trap is
          backflushed while being rapidly heated in order to thermally
          desorb the components into the inlet of a gas chromatograph.  The
          components are separated via the gas chromatograph and detected
          using a mass spectrometer which is used to provide both
          qualitative and quantitative information.  The chromatographic
          conditions as well as typical mass spectrometer operating
          parameters are supplied.
3.  Interferences
     3.1  Interferences coextracted from the samples will vary considerably
          from source to source, depending upon the diversity of the
          industrial complex or municipality being sampled.  Impurities in
          the purge gas and orgaric compounds out-gas ing from the plumbing

-------
          ahead of the trap account for the majority of contamination
          problems.   The analytical system must be demonstrated to be free
          from interferences under the conditions of the analysis by running
          method blanks.  Method  blanks are run by charging the purging
          device with organic-free water and analyzing it in a normal
          manner.  The use of non-TFE plastic tubing,  non-TFE thread
          sealants,  or flow controllers with rubber components in the
          purging device should be avoided.
     3.2  Samples can be contaminated by diffusion of volatile organics
          (particularly methylene chloride) through the septum seal into the
          sample during shipment  and storage.  A field blank prepared from
          organic-free water and  carried through the sampling and handling
          protocol can serve as a check on such contamination.
     3.3  Cross contamination can occur whenever high level and low level
          samples are sequentially analyzed.  To reduce cross contamination,
          it is recommended that   the purging device and sample syringe be
          rinsed out twice, between samples, with organic-free water.
          Whenever an unusually concentrated sample is encountered, it
          should be followed by an analysis of organic-free water to check
          for cross-contam- ination.  For samples containing large amounts
          of water soluble materials, suspended solids, high boiling
          compounds or high organohalide levels it may be necessary to wash
          out the purging device with a soap solution, rinse with distilled
          water, and then dry in a 105°C oven between analyses.
4.  Apparatus and Materials
     4.1  Sampling equipment, for  discrete  sampling.

-------
4.1.1       Vial, with cap - 40 ml capacity screw cap  (Pierce #13075 or
            equivalent).  Detergent wash and dry vial  at 105°C for
            one hour before use.
     4.1.2  Septum - Teflon-faced silicone (Pierce #12722 or
            equivalent).  Detergent wash and dry at 105°C for one
            hour before use.
4.2  Purge and trap device - The purge and trap equipment consists of
     three separate pieces of apparatus:  a purging device, a trap, and
     a desorber.  The complete device is available commercially from
     several vendors or can be constructed in the laboratory according
     to the specifications of Bellar and Lichtenberg (Ref.  2,3).  The
     sorbent trap consists of 1/8 in. O.D.  (0.105 in.  I.D.) x 25 cm
     long stainless steel tubing packed with 25 cm of Tenax-GC (60-80
     mesh) and 8 cm of Davison Type-15 silica gel (35-60 mesh).   See
     Figures 1 through 4.  Ten centimeters traps may be used providing
     that the recoveries are comparable to the 25 on traps.
4.3  Gas chromatograph - Analytical system complete with a temperature
     programmable gas chromatograph suitable for on-column  injection
     and all required accessories including an analytical column.
     4.3.1  Column 1 - An 8 ft.  stainless steel column (1/8 in.  00 0.90
            to 0.105 in. ID) packed with 1% SP-1000 coated on 60/80
            mesh Carbopack B preceded by a 5 cm precolumn packed with
            1% SP-1000 coated on 60/80 mesh Chromosorb W.  A glass
            column (1/4 in OD x  2 mm ID) may be substituted.  The
            precolumn is necessary only during conditioning.
                        IF

-------
     4.3.2  Column 2 - An 8 ft. stainless steel column (1/8 in OD x
            0.09 to 0.105 in. ID) packed with 0.2% Carbowax 1500 coated
            on 60/80 mesh Carbopack C preceded by a 1 ft. stainless
            steel column (1/8 in. OD x 0.09 to 0.105 in. ID) packed
            with 3% Carbowax 1500 coated on 60/80 mesh Chromosorb W.  A
            glass column (1/4 in. OD x 2 mm ID) may be substituted.
            The precolumn is necessary only during conditioning.
4.4  Syringes - glass 5-ml hypodermic with Luer-Lok tip  (3 each).
4.5  Micro syringes - 10, 25, 100 ul.
4.6  2 -way syringe valve with Luer ends (3 each, Teflon  or Kel-F).
4.7  Syringe - 5-ml gas-tight with shut-off valve.
4.8  8-inch, 20-gauge syringe needle - One per each 5-ml syringe.
4.9  Mass Spectrometer - capable of scanning from 20-260 in six seconds
     or less at 70 volts (nominal), and producing a recognizable mass
     spectrum (background corrected) at unit resolution  from 20 ng of
     4-bromofluorobenzene (BFB) when injected through the GC inlet.
     The mass spectrometer must be interfaced with a gas chromatograph
     equipped with an injector system designed for all -glass on-column
     packed column work.  All sections of the transfer lines must be
     glass or glass-lined and deactivated  Use Sylon-CT, Supelco, (or
     equivalent) to deactivate.  The GC/MS interface can utilize any
     separator that gives recognizable mass spectra  (background
     corrected) and acceptable calibration points at the limit of
     detection specified for each compound in Table  II.
4.10 A computer system should be interfaced to the mass  spectrometer to
     allow acquisition of continuous mass scans for the  duration of the
                         IF

-------
          chromatographic program.  The computer system should also be
          equipped with mass storage devices for saving all data from GC-MS
          runs.   There should be computer software available to allow
          searching any GC/MS run for specific ions and plotting the
          intensity of the ions with respect to time or scan number.  The
          ability to integrate the area under an specific ion plot peak is
          essential for quantification.
5.  Reagents
     5.1  Sodium thiosulfate - (ACS) Granular.
     5.2  Trap Materials
          5.2.1   Porous polymer packing 60/80 mesh chromatographic grade
                 Tenax GC (2,6-diphenylene oxide).
          5.2.2   Three percent OV-1 on Chromosorb-W 60/80 mesh.
          5.2.3   Silica gel-(35-60 mesh) .- Davison, grade-15 or equivalent.
     5.3  Activated carbon - Filtrasorb-200 (Calgon Corp.) or equivalent.
     5.4  Organic-free water
          5.4.1   Organic-free water is defined as water free of interference
                 when employed in the purge and trap procedure described
                 herein.  It is generated by passing tap water or well water
                 through a carbon filter bed containing about 1 Ib. of
                 activated carbon.
          5.4.2   A water system (Millipore Super-Q or equivalent) may be
                 used to generate organic-free deionized water.
          5.4.3   Organic-free water may also be prepared by boiling water
                 for 15 minutes.  Subsequently, while maintaining the
                 temperature at 90°C, bubble a contaminant-free inert gas

-------
            through the water for one hour.  While still hot, transfer
            the water to a narrow mouth screw cap bottle equipped with
            a Teflon seal.
5.5  Stock standards (2 mg/ml) - Prepare stock standard solutions in
     methanol using assayed liquids or gases as appropriate.  Because
     of the toxicity of some of the organohalides, primary dilutions of
     these materials should be prepared in a hood.  A NIOSH/MESA
     approved toxic gas respirator should be worn when the analyst
     handles high concentrations of such materials.
     5.5.1  Place about 9.8 ml of methanol into a 10 ml ground glass
            stoppered volumetric flask.  Allow the flask to stand,
            unstoppered, for about 10 minutes or until all alcohol
            wetted surfaces have dried.  Tare the flask to the nearest
            0.1 mg.
     5.5.2  Add the assayed reference material:
            5.5.2.1  Liquids - Using a 100 ul syringe, immediately  add
                     2 to 3 drops of assayed reference material to  the
                     flask, then reweigh.  Be sure that the drops fall
                     directly into the alcohol without contacting the
                     neck of the flask.
            5.5.2.2  Gases - To prepare standards of bromomethane,
                     chloroethane, chloromethane, and vinyl chloride,
                     fill a 5 ml valved gas-tight syringe with the
                     reference standard to the 5.0--1 mark.   Lower  the
                     needle to 5 mm above  the re:\-r"  alcohol  menicus.
                     Slowly inject the reference  r,  -,-srd into the  neck

-------
                     of flask (the heavy gas will rapidly dissolve into
                     the methyl alcohol).
     5.5.3  Reweigh the flask, dilute to volume, stopper, then mix by
            inverting the flask several times.  Transfer the standard
            solution to a 15 ml screw-cap bottle equipped with- a Teflon
            cap liner.
     5.5.4  Calculate the concentration in mg per ml (equivalent to ug
            per ul) from the net gain in weight.
     5.5.5  Store stock standards at 4°C.  Prepare fresh standards
            every second day for the four gases and 2-chloroethylvinyl
            ether.  All other standards must be replaced with fresh
            standards each week.  Remove samples from cold storage
            (approximately an hour prior to analysis) and bring to room
            temperature by placing in a warm water bath at 20-25°C.
5.6  Surrogate Standard Dosing Solution - From stock standard solutions
     prepared as above, add a volume to give 1000 ug each of
     bromochloromethane, 2-bromo-l-chloropropane, and
     1,4-dichlorobutane to 40 ml of organic-free water contained in a
     50 ml volumetric flask, mix and dilute to volume.  Prepare a fresh
     surrogate standard dosing solution weekly.  Dose the surrogate
     standard mixture into every 5 ml sample and reference standard
     analyzed.
Calibration
6.1  Using stock standards, prepare secondary dilution standards in
     methanol that contain the compounds of interest, either singly or
     mixed together.  The str :--ds should be at concentrations such

-------
     that the aqueous standards prepared in 6.2 will completely bracket
     the working range of the chroma to graphic system.  If the limit of
     detection listed in Table 2 is 10 ug/1, for example, prepare
     secondary methanolic standards at 100 ug/1, and 500 ug/1, so that
     aqueous standards prepared from these secondary calibration
     standards, and the primary standards, will define the linearity of
     the detector in the working range.
6.2  Using both the primary and secondary dilution standards, prepare
     calibration standards by carefully adding 20.0 ul of the standard
     in methanol to 100, 500, or 1000 ml of organic-free water.  A 25
     ul syringe (Hamilton 702N or equivalent) should be used for this
     operation.  These aqueous standards must be prepared fresh daily.
6.3  Assemble the necessary gas chromatographic apparatus and establish
     operating parameters equivalent to those indicated in Table 2.  By
     injecting secondary dilution standards, establish the linear range
     of the analytical system for each compound and demonstrate that
     the analytical system meets the limits of detection requirements
     in Table 2.
6.4  Assemble the necessary purge and trap device.  Pack the trap as
     shown in Figure 2 and condition overnight at a nominal 180°C by
     backflushing with an inert gas flow of at least 20 ml/min.  Prior
     to use, daily condition the traps for 10 minutes by backflushing
     at 180°C.  Analyze aqueous calibration standards (6.2) according
     to the purge and trap procedure in Section 9.  Compare the
     responses to those obtained by injection of standards (6.3), to
     determine the analytical precision.  The analytical precision of
                              3.13

-------
     the analysis of aqueous standards must be comparable to data
     presented by Bellar arid Lichtenberg (1978, Ref. 1) before reliable
     sample analysis may begin.
6.5  The internal standard approach is acceptable for the purgeable
     organics when internal standards meeting the following criteria
     can be selected:
     6.5.1  No interference with other components in the sample.
     6.5.2  Structural  similarity to the compound to be determined.
     6.5.3  Elutes close to the compound to be determined.
     6.5.4  The amount added approximates the concentration of the
            compound to be determined.
6.6  Internal Standard Method - The utilization of the internal
     standard method requires the periodic determination of response
     factors (RF) which are defined in equation 1.
       Eq. (1)  RF = (ASC1S)/(A.SCS)
     Where:
          A  is the integrated area or peak height of the
          characteristic ion for the priority pollutant standard.
          A.  is the integrated area or peak height of the
          characteristic ion for the internal standard.
          C.  is the amount of the internal standard in ug.
          C  is the amount of the priority pollutant standard in ug.
     The relative response ratio for the priority pollutants should be
     known for at least two concentration values--50 ng injected to
     approximate 10 ug/1 and 500 ng to approximate the 100 ug/1 level.
     Those compounds that do not respond at either of these levels may

-------
     be run at concentrations appropriate to their response.  The
     response factor (RF) must be determined over all concentration
     ranges of standard (C ) which are being determined.  (Generally,
     the amount of internal standard added to each extract is the same
     so that C.  remains constant.)  This should be done by preparing
     a calibration curve where the response factor (RF) is plotted
     against the standard concentration (C ).  Use a minimum of three
     concentrations over the range of interest.  Once this calibration
     curve has been determined, it should be verified daily by
     injecting at least one standard solution containing internal
     standard.  If significant drift has occurred, a new calibration
     curve must be constructed.
     Note:  EPA, through its contractors and certain of its Regional
            Laboratories, is currently evaluating selected compounds
            for use as internal standards in the analysis of organics
            by purge and trap.
6.7  The external standard method can be used at the discretion of the
     analyst and is recommended when the criteria for use of the
     internal standard cannot be met.  Prepare a master calibration
     curve using a minimum of three standard solutions of each of the
     compounds that are to be measured.  Plot concentrations versus
     integrated areas or peak heights (selected characteristic ion for
     GC/MS).  One point on each curve should approach the method
     detection limit. After the raster set of instrument calibration
     curves have been establishes, they should be verified daily by
     injecting at least one standard solution.  If significant drift
     has occurred, a new calibration curve must be

-------
          constructed.
7.   Quality Control
     7.1  Before processing any samples, the analyst should daily
          demonstrate, through the analysis of an organic-free water method
          blank, that the entire analytical system is  interference-free.
     7.2  Standard quality assurance practices should be used with this
          method.  Field replicates should be collected to validate the
          precision of the sampling .technique.  Laboratory replicates should
          be analyzed to validate the precision of the analysis.  Fortified
          samples should be analyzed to validate the accuracy of the
          analysis.
     7.3  Identification of a peak on the chromatogram is done by analysis
          of the entire background corrected mass spectrum.
     7.4  The analyst should maintain constant surveillance of both the
          performance of the analytical system and the effectiveness of the
          method in dealing with each sample matrix by spiking each 5 ml
          sample, standard, and blank with surrogate halocarbons.  A
          combination of bromochloromethane, 2-bromo-l-chloropropane, and
          1,4-dichlorobutane is recommended to encompass the boiling range
          of compounds covered by this method.  Prepare a fresh -surrogate
          standard dosing solution on a weekly basis.
8.   Sample Collection, Preservation, and Handling
     8.1  Grab samples must be collected in glass containers having a total
          volume greater than 20 ml.  Fill the sample bottles in such a
          manner that no air bubbles pass through the sample as the bottle
          is being filled.  Seal the bottles so that no air bubbles are

-------
          entrapped in it.   Maintain the hermetic seal on the sample bottle
          until  time of analysis.
     8.2  The sample must be iced or refrigerated from the time of
          collection until  extraction.  If the sample contains residual
          chlorine, add sodium thiosulfate preservative (10 mg/40 ml) to the
          empty sample bottles just prior to shipping to the sample site,
          fill with sample just to overflowing, seal the bottle, and shake
          vigorously for 1  minute.
     8.3  All samples must be analyzed within 7 days of collection.
9.   Sample Extraction and Gas Chromatography
     9.1  Adjust the purge gas (nitrogen or helium) flow rate to 40 ml/min.
          Attach the trap inlet to the purging device, and set the device to
          the purge mode.  Open the syringe valve located on the purging
          device sample introduction needle.
     9.2  Remove the plunger from a 5 ml syringe and attach a closed syringe
          valve.  Open the sample bottle (or standard) and carefully pour
          the sample into the syringe barrel until  it overflows.  Replace
          the syringe plunger and compress the sample.  Open the syringe
          valve and vent any residual air while adjusting the sample volume
          to 5.0 ml.  Since this process of taking  an aliquot destroys the
          validity of the sample for future analysis, the analyst should
          fill a second syringe at this time to protect against possible
          loss of data.  Add 5.0 ul of the surrogate spiking solution  (7.3)
          through the valve bore, then close the valve.
     9.3  Attach the syringe-valve assembly to the  syringe valve on  the
          purging device.  Open the syringe valve and inject the sample  into

-------
     the purging chamber.
9.4  Close both valves and purge the sample for 12.0 + .05 minutes.
9.5  After the 12-minute purge time, attach the trap to the
     chromatograph, and adjust the device to the desorb mode.
     Introduce the trapped materials to the GC column by rapidly
     heating the trap to 180°C while backflushing the trap, with an
     inert gas, at 20 to 60 ml/min for 4 minutes.  If rapid heating
     cannot be achieved, the gas chromatographic column must be used as
     a secondary trap by cooling it to 30°C (or subambient, if
     problems persist) instead of the initial program temperature of
     45°C.
9.6  While the trap is being desorbed into the gas chromatograph, empty
     the purging chamber using the sample introduction syringe.  Wash
     the chamber with two 5 ml flushes of organic-free water.  After
     the purging device has been emptied, continue to allow the purge
     gas to vent through the chamber until the frit is dry, and ready
     for the next sample.
9.7  After desorbing the sample for four minutes recondition the trap
     by returning the purge and trap device to the purge mode.  Wait 15
     seconds then close the syringe valve on the purging device to
     begin gas flow through the trap.  Maintain the trap temperature at
     180°C.  After approximately seven minutes, turn off the trap
     heater and open the syringe valve to stop the gas flow through the
     trap.  When cool, the trap is ready for the next samp I.e.  (Note:
     If this bake out step is omitted, the amount of water entering the
     GC/MS system will progressively increase causing deterioration of
                        or

-------
          and potential  shut  down of the system.)
     9.8  The analysis of blanks is most important in the purge and trap
          technique since the purging device and the trap can be
          contaminated by residues from very concentrated samples or by
          vapors in the laboratory.  Prepare blanks by filling a sample
          bottle with organic-free water that has been prepared by passing
          distilled water through a pretested activated carbon column.
          Blanks should be sealed, stored at 4°C,  and analyzed with each
          group of samples.
10.  Gas Chromatography - Mass Spectrometry
     10.1  Table 2 summarizes  the recommended gas chromatographic column
          materials and operating conditions for the instrument.  Included
          in this table are estimated retention times and sensitivities that
           /
          should be achieved  by this method.  An example of the separation
          achieved by Column  1 is shown in Figure 5.
     10.2  GC-MS Determination - Suggested analytical conditions for
          determination of the priority pollutants .amenable to purge and
          trap, using the Tekmar LCS-1 and a GC/MS are given below.
          Operating conditions vary from one system to another; therefore,
          each analyst must optimize the conditions for each purge and trap
          and GC/MS system.
     10.3  Purge Parameters
          Sample size - 5.0 ml.
          Purge gas - Helium, high purity grade
          Purge time - 12 minutes
          Purge flow - 40 ml/min.
                           jr

-------
          Trap Dimensions - -1/8 in. O.D. (0.105 in. I.D.) x 25 cm long
          Trap sorbent -  Tenax-GC, 60/80 mesh (15 an], plus Type 15 silica
                          gel, 35/60 mesh (8 cm).
          Desorption flow - 20 ml/min.
          Desorption time - 4 min.
          Desorption temperature - 180°C
     10.4 Mass Spectrometer Parameters
          Electron energy - 70 volts (nominal)
          Mass range - 20-27, 33-260 amu
          Scan time - 6 seconds or less.
     10.5 Calibration of the gas chromatography-mass spectrometry (GC-MS
          system - Evaluate the system performance each day that it is to be
          used for the analysis of samples  or blanks by injecting 20 ng of
          BFB into the GC inlet.
     10.6 Analyze the 20 ug/1 standard to demonstrate instrument performance
          or these compounds.
11.   Qualitative and Quantitative Determination
     11.1 The complete background corrected mass  spectrum is compared to a
          reference mass spectrum, either from an authentic sample or a
          library spectrum,  to provide qualitative identification.   All  of
          the following criteria must be met.
          11.1.1  The experimental  mass spectrum must contain all  ions
                 present above 1% relative  abundance in the mass spectrum of
                 the reference with the general agreement of _+ 20% of the
                 relative abundance in the  reference mass spectrum.
          11.1.2 Ions present in the experimental  mass spectrum that are not

-------
            present in the reference mass spectrum must not exceed 10X
            of the total  ion abundance in the experimental mass
            spectrun.   These contaminant ions should be readily
            accounted- for by common background Ions.  If the
            experimental  mass spectrun 1s highly contaminated,
            Identificaton should not be accepted.
     11.1.3 The retention time at the experimental mass spectrun should
            be within  £ 5 spectrum numbers or * 30 seconds (whichever
            Is greater) of the retention time of the authentic compound.
     11.1.4 Structural Isomers that have very similar mass spectra
            (phenanthrene, anthracene; benzola)pyrene, benzo(e)pyrene;
            etc.) can  be explicitly Identified if the resolution
            between the Isomers in a standard mix is acceptable.
            Acceptable resolution 1s achieved if the valley height
                                         ZS'L'f
            between Isomers is less than be44 the sum of the two peak
            heights.  Otherwise, structural Isomers are identified as
            1 saneric pairs.
11.2 The primary Ions  to be used to quantify each compound are listed
     in Table 3.  If the sample produces an interference for the
     primary ion, use  a secondary ion to quantify.
11.3 For low concentrations, or direct aqueous injection of
     acrylonltrile and acrolein, the characteristic masses (Table 3)
     may be used for selected ion monitoring (SIM).  SIM is the use of
     a mass spectrometer as a substance selective detector by measuring
     the mass spectrometric response at one or several characteristic
     masses in real time.

-------
11.4 Internal  Standard Method Calculations - By adding a constant known
     amount of internal standard (C.  in ug) to every sample extract,
     the concentration of priority pollutant (C ) in ug/1 in the
     sample is calculated using equation 2.
                                    (A5) (C.s)
Eq<  (2)  C
                          o  =    (ASJ (RP) (Vo)
     Where:
           V  is the volume of the original sample in liters.
     To quantify, add the internal standard to the 5.0 ml sample no
     more than a few minutes before purging to minimize the possibility
     of losses due to evaporation, adsorption, or chemical reaction.
     Calculate the concentration by using the previous equations with
     the appropriate response factor taken from the calibration curve.
     (Ideally, the response factor will  not change with concentration.)
11.5 External  Standard Method Calculations - The concentration of the
     unknown can be calculated from the  slope and intercept of the
     curve.   The unknown concentration can be determined using equation
     3.
                                                   (A)(V )
         Eq.  (3)     Micrograms/liter = ng/ml  =
      Where:
           A  - mass of compound from calibration curve (ng).
           V.  * volume of extract injected (ul)
           Vt  = volume of total extract (ul)
           V   = volume of water extracted (ml)

-------
     11.6  An alternate external  standard approach for purgeables utilizes a
           single point calibration.  Prepare and analyze a reference
           standard that closely approximates the response for each
           component in a sample.  Calculate the concentration in the sample
           using Equation 4.
          Eq. 4    micrograms per liter  =

          where:  A is the area of the unknown
                  B is the concentration of standard
                  C is the area of the standard
    11.7  Report all results  to two significant figures.  Report internal
          standard data to two significant figures-.  When duplicate and
          spiked samples are  analyzed, all data obtained should be
          reported.  Report results in micrograms per liter without
          correction for recovery data.  Include identification of any
          components found in the blanks.
12. Precision and Accuracy
    12.1  Determine the precision of the method by dosing blank water with
          the compounds selected as surrogate standards -
          bromochloromethane, 2-bromo-l-chloropropane, and
          1,4-dichlorobutane  - and running replicate analyses.  Calculate
          the recovery and its standard deviation.  These compounds
          represent early, middle, and late eluters over the range of the
          Consent Decree compounds and are not, themselves, included on  the
          list.

-------
    12.2 The sample matrix can affect the purging efficiencies of Individual
         compounds; therefore, each sample must be dosed with the surrogate
         standards and analyzed 1n a manner Identical to the Internal
         standards in blank water.  If the recovery of the surrogate
         standard shows a-deviation greater than two standard deviations
         (12.1), repeat the dosed sample analyses.  If the deviation 1s
         again greater than two standard deviations, dose another aliquot of
         the same sample with the compounds of Interest at approximately two
         times the measured values and analyze.  Calculate the recovery for
        . the Individual compounds using these data.
13. References
    1.   "The Analysis of Halogenated Chemical Indicators of Industrial
         Contamination 1n Water by the Purge and Trap Method," U.S. EPA,
         Environmental Monitoring and Support Laboratory, Cincinnati, OH,
         45268, Dec. 1978.
    2.   "Symposium on Measurement of Organic Pollutants in Water and
         Wastewater," ASTM Special Publication, 1979 (In Press).
    3.   "Determining Volatile Organics at Microgram-per-Litar Levels by Gas
       •  Chromatography," T. A. Bellar and J. J. Lichtenberg, Jour. AWWA,
         66, 739-744, Dec. 1974.
    4.   ASTM Annual Standards - Water, part 31, Method D2908 "Standard
         Recommended Practice for Measuring Water by Aqueous-Injection Gas
         Chromatography."
    5.   ASTM Annual Standards - Water, part 31, Method 03371 "Tentative
         Method of Test for Nitriles in Aqueous Solution of Gas Liquid
         Chromatograph."

-------
6.   "Direct Analysis  of  Water  Samples for Organic Pollutants with Gas
     Chromatography-Mass  Spectrometry," Harris, L. E., Budde, W. L., and
     Eichelberger,  J.  W.  Anal.  Chem., 4§, 1912  (1974).
                       3T-

-------
                          Table  1
Parameter

Acrolein
Acrylonitrile
Benzene
Bromomethane
Bromodi chloromethane
Bromoform
Carbon Tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethylvinyl ether
Chloroform
Chloromethane
Dibromoch1oromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-1,2-Dichloroethene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-1,3-Dichloropropene
Ethylbenzene
Methylene chloride
1,1,2,2-Tetrachloroethane
Tetrachlorethene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Trichlorofluoromethane
Toluene
Vinyl chloride
STORET No.

  34210
  34215
  34236
  34413
  32101
  32104
  32102
  34301
  34311
  34576
  32106
  34418
  34105
  34496
  34531
  34501
  34546
  34541
  34561
  34561
  34371
  34423
  34516
  34475
  34506
  34511
  39180
  34488
  34010
  39175

-------
                                   TABLE 2

              Gas  Chromatography  of  Organlcs by Purge and Trap
Compound
chlororaethane
bromooiethane
vinyl chloride
chloroethane
methylene chloride
tr 1 ch 1 orofl uoromethane
1,1-d1chloroethene
bromochloromethane (SS)
1 , 1 -d1 ch 1 oroethane
trans- 1 ,2-d1chloroethene
nhl T.LI +f iLiin
cn i wru i u< in
1 ,2-d1chloroethane
1,1,1 -tr1 chloroethane
carbon tetrachlorlde
bromodl ch 1 oromet hane
1 ,2-d1cn1oropropane
• 	 ^ 4k I • I 4
trans- 1 ,j-
-------
Column 2 Eight ft. stainless steel column (1/8 1n. OD x 0.1 in. ID) packed
         with 0.2% Carbowax 1500 coated on 60/80 mesh Carbopack C preceded
         by a 1 ft. stainless steel column (1/8 in. OD x 0.1 in. ID) packed
         with 3% Carbowax 1500 coated on 60/80 mesh Chromosorb W.  A glass
         column (1/4 in. OD x 2 mm ID) may be substituted.  Carrier gas:
         helium at 40 ml/min.  Temperature program:  3 min. isothermal at
         60°C then 8°/min to 160°, hold at 160 until all compounds
         elute.

* This is a minimum level at which the entire system must give recognizable
  mass spectra (background corrected) and acceptable calibration points.

** Sensivitity refers to either qualitative screening with this method or
   direct aqueous injection GC-FID (Ref. 4,5,6).

-------
                                   TABLE 3
Compound
                   Characteristic Ions of Volatile Organlcs

                                 El Ions
chloromethane
bromomethane
vinyl chloride
chloroethane
methyl ene chloride
tr1ch1orofluoromethane
I,1-d1chloroethene
bromochlorcmethane (IS)
1,1 -d1 chloroethane

trans- 1,2-dichloroethene
chloroform
1,2-di chloroethane
1,1,1-trichloroethane
carbon tetrachlorlde
bromodl chloromethane
l,2-d1chloropropane
trans-l,3-
-------
   OPTIONAL
   FOAM
   TRAP
•BUT'/« IN.
     0. D.
                  —14MM 0. 0.
                   INLET % IN.
                        0. D.
V4 IN.
0. D. EXIT
             11
             11
             I'
           INLET

    2-WAY SYRINGE VALVE
    17CM. 20 GAUGE SYRINGE NEEDLE

       . 0. 0. RUBBER SEPTUM
   10MM GLASS FRIT
   MEDIUM POROSITY
                        _10IHM. 0. D.
                           INLET
                           '/«IN. 0. 0.
                     1/T6 IN. O.D.
                     STAINLESS STEEL
                                       13X MOLECULAR
                                       SIEVE PURGE
                                       GAS FILTER
I                                          PURGE GAS
                                          FLOW
                                          CONTROL
             Rgure 1. Purging device

-------
LJJ
CC
                                                       03
                                                       U
                                                       c
                                                       c
                                                       o
                                                       O
                                                       O

                                                      •O

                                                       s.?
                                                       en ~
                                                       05-C
                                                       c  ^
O
O
cc
Q.

O
o
<
a.

I   ^
o J£ S

§ yj <
                           CJ
                           LD
      to



      CN

      O
00
          a
                         i- b

-------








LU
O
S
COLUMN
1 z
3 1
u g
». <->
ce
0 =i

s \ '/>








t/i
<
a
ee
UJ
i
<
(j

O
a.
z
0
^
CJ
LU
i^
a
g
_j
^i*t v^l «^
id.
5^
ci:E
-."^.i.
vT^P f"o~!
T_U- L_^
PCS-^i
1 ^r^r
"^^



2
S
3
0
CJ
H»
oc
O
a.

—i
^
g
^*
0.
O
/

r=.




LU
>
<
>
Z
O
u
LU
_i
LU
uo


^r_


EC
H-
<
LU
X
LU
oc
S
LU
U

, , rf
LU 4
_J I—
Z <£
75
a. LU-
< a:
"^

j
g
i—
o
u
\
\ 	
/ u.
J u.
P''s2
, 2=
1 <
cc
>—
\ 1
r~T"
\~5
h-LU \ ^
s>
O_i
°r<
05 >
	 =




sj
0
0
C«J
CM
Si1
P i:
S
UJ

h—
LU
OQ
0)
LU
13
_<
<







u
c
c
<
a.
<
ce
H~
a
z
5
ee
a.

\g>-fc S
0 °°
z o
to H—

LU
a
>
LU
a
i
i

f
">=*=1N3A '
~^*=!'


;i
rj — -I ^
_, , A >A,
0 — ,. ./ — : i
cc ^
t— •
2 ^
0
5j
S
_J


-' •f^==
f ' '
I
oe 'i
. O '1









	



U. Uj ,_ .i
£S <
"a
U1 J


U
fi
J CJ
? LJ i




£ QC













-=k

~s~~
_J
o
ac
LO 5
< 5
u F,
LU U

C S
§3
a. u.






^=§^^
I








K>
<
= 55
u t=
LU _J
o"-
S i^
>
S S
«— trt






























i

— 1










O
E
o
OJ
3
a
i
device
a
CO
k.
c

-------
         CARRIER GAS
         FLOW CONTROL  LIQUID INJECTION PORTS
   PRESSURE
   REGULATOR
                                          COLUMN OVEN
PURGE GAS  ,  _
FLOW CONTROL
  13X MOLECULAR
  SIEVE FILTER
  UT.L n	l'	CONFIRMATORY COLUMN
    S| ^-TVVJ^TO DETECTOR
    £i|	••'-  j  ^-ANALYTICAL COLUMN
    v  OPTIONAL 4-PORT COLUMN
      SELECTION VALVE
!6-PORT TRAP INLET
                                                 CONTROL
                                  PURGING
                                  DEVICE
                     Note:
                     ALL LINES BETWEEN
                     TRAP AND GC
                     SHOULD BE HEATED
                                             TO 80°C.

Figure 4. Schematic of purge and trap device - desorb mode

-------
  o
  a
  O
  CM
  CM
CO GO
O 5
03 \
r-
a. 6                           3N3ZN3aOIAIOaa
oo a

ic ^
«—

I I                       3NVX3HOa01HD-L

= C                    3IM32M3aOa01H3

gg
       3MWLJl3OU(^TLJ'^HUI3!-7  *7 'I *l
       ji\JVr1ijUaw Irl*J vuJ.JJ. C  6 L  L
                                ^i^

           3NVdoadoaoiH3iai-e 7'


                    ^MHMMMMMM

31MVH130a01HOVai3i-3' I' L' L

                       3NVHi301AIOaaiQ-2  'I

            3M3dOadOa01HOIQ-C 'I

        3N3dOadOa01HOIQ-C 'L -
       3IMVdOadOa01HOIQ-2 '
       3I\IVHi30a01HDiai -I'L'L
      3IM3H130a01HOIQ-L'L
                                                   oo
                                                   CNJ
                                                   CM
                                                   CM  -   =
                                                   0,2
                                                   CM
                                                   00
u
'E
                                                          J2
                                                          o
                                                          >
                                                          •»—
                                                          O

                                                          E
                                                          re
                                                          w
                                                          CT
                                                          O

                                                          ra
                                                          £

                                                          O
                                                          w

                                                          "o

                                                          (/>
                                                          re
                                                          O


                                                          if)

                                                          09
                                                          ha
                                                          •tf

                                                          S5
                                                          iZ

-------
                     BASE/NEUTRALS, ACIDS, AND PESTICIDES
                                  METHOD 625
1.  Scope and Application
    1.1   This method covers the determination of those "unambiguous priority
         pollutants" associated with the Consent Decree that are solvent
         extractable and amenable to gas chromatography.  The parameters
         listed in Tables 1 and 2 may be determined by this method.  This
         method is qualitative for the parameters listed in Table 3.
    1.2   This method is applicable to the determination of these compounds
         in municipal and industrial discharges.  It is designed to be used
         to meet the survey requirements of the Effluent Guidelines Division
         (EGD) and the monitoring requirements of the National Pollutants
         Discharge Elimination System (NPDES).
    1.3   The detection limit of this method is usually dependent upon the
         level of interferences rather than instrumental limitations.  The
         limits listed in Tables 4, 5, and 6 represent levels that can be
         achieved in wastewaters, as such they were determined by the
         minimum quantity that must be injected into the system to get
         confirmation by the mass spectrometric method described below.
    1.4   The GC/MS parts of this method are recommended for use only by
         analysts experienced with GC/MS or under the close supervision of
         such qualified persons.
2.  Summary of Method
    2.1   A 1-liter sample of wastewater is extracted, with methylene
                             JI

-------
         chloride using separatory funnel or continuous extraction
         techniques.  If emulsions are a problem, continuous extraction
         techniques should be used.  The extract 1s dried over sodium
         sulfate and concentrated to a volume of 1 ml or less using a
         Kuderna-Oanish (K-0) evaporator.  Chroraatographic conditions are
         described which allow for the separation of the compounds In the
         extract.
    2.2  Identification of a compound (qualitative analysis) is performed by
         an analysis of the full, background corrected, mass spectra.
    2.3  Quantitative analysis is performed'by GC/MS using either the
         internal standard or external standard technique.  (Base/Neutral,
         and.Acids).  Quantitative analysis of the pesticides is done using
         electron capture gas chroaatography (Federal Register, 38, Vol.
         125, Part II, pp. 17318-17323, Friday, June 29, 1973).  One liter
         rather than 100 ml is to be extracted.
3.  Interferences
    3.1  Solvents, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of chromatograras.  All of these materials must be
         demonstrated to be free from interferences under the conditions of
         the analysis by running method blanks.  Specific selection of
         reagents and. purification of solvents by distillation in all-glass
         systems may be required.
    3.2  Interferences coextracted from the samples will vary considerably
         from source to source, depending upon the diversity of the
         industrial complex or municipality being sampled.

-------
    3.3  The recommended analytical procedure may not have sufficient
         resolution to differentiate between certain 1somer1c pairs.  These
         are anthracene and phenanthrene, chrysene and benzo(a)anthracene,
         and benzo(b)fluoranthene and benzo(k)f1uoranthene.  The GC
         retention time and mass spectral data are not sufficiently unique
         to make an unambiguous distinction between these compounds.
         Alternative techniques should be used -to Identify these specific
         compounds.  See Reference 1.
4.  Apparatus and Materials
    4.1  Sampling equipment, for discrete or composite sampling.
         4.1.1  Grab sample bottle - amber glass, 1-liter or 1-quart
                volume.  French or Boston Round design is recommended.  The
                container must be washed.and solvent rinsed before use to
                minimize interferences.
         4.1.2  Bottle caps - Threaded to fit sample bottles.  Caps must be
                lined with Teflon.  Aluminum foil may be substituted if
                sample 1s not corrosive.
         4.1.3  Compositing equipment - Automatic or manual compositing
                system.  Must incorporate glass sample containers for the
                collection of a minimum of 1000 ml.  Sample containers must
                be kept refrigerated during sampling.  No plastic or rubber
                tubing other than. Teflon may be used in the system.
    4.2  Separatory funnel - 2000 ml, with Teflon stopcock  (Aca Glass
         7228-T-72 or equivalent).
    4.3  Drying column - A 20 mm ID pyrex cnromatograpnic  column equipped
         with coarse glass frit or glass wool plug.

-------
4.4  Kudema-Danlsh (K-0) Apparatus
     4.4.1  Concentrator tube - 10 ml, graduated (Kontes K- 57 0050- 1025
            or equivalent).  Calibration must be checked.  Ground glass
            stopper (size 19/22 Joint) 1s used to prevent evaporation of
            extracts.
     4.4.2  Evaporative flask - 500 ml (Kontes K-5700T-0500 or
            equivalent).  Attach to concentrator tube with springs.
            (Kontes K-662750-0012).
     4.4.3  Snyder column - three-ball macro (Kontes KS03000-0232 or
            equivalent).
     4.4.4  Snyder column - two-ball micro (Kontes K-569002-0219 or
            equivalent).
     4.4.5  Boiling chips-extracted, approximately 10/40 mesh.
4.5  Water bath - Heated, with concentric ring cover, capable of
     temperature control (+ 2°C).  The bath should be used 1n a hood.
4.6  Gas chromatograph - Analytical system complete with gas
     chromatograph capable of on-column Injection and all required
     accessories including column supplies, gases, etc.
     4.6.1  Column 1 - For Base/Neutral and Pesticides a 6-foot glass
            column (1/4 in 00 x 2 mm 10) packed with 32 SP-2250 coated
            on 100/120 Supelcoport (or equivalent).
     4.6.2  Column 2 - For Adds, a 6-foot glass column (1/4 in 00 x 2
            mm ID) packed with IS SP-1240 OA coated on 100/120 mesh
            Supelcoport (or equivalent).
4.7  Mass Spectrometer - Capable of scanning from 35 to 350 a.m.u.
     (Base/Neutrals-Acids) or 35 to 450 a.m.u. (Pesticides) every 6
                    IT-

-------
     seconds or less at 70  volts (nominal)  and producing a recognizable
     mass spectrum (background corrected)  at unit resolution from 50 ng
     of OFTPP when the sample 1s Introduced through the GC Inlet
     (Reference 1).  The mass spectrometer must be interfaced with a gas
     chromatograph equipped with an Injector system designed for
     splltless Injection glass capillary column use or an Injector
     system designed for all -glass on-column packed column use.  All
     sections of the transfer lines must be glass or glass-lined and
     must be deactivated.  (Use Sylon-CT,  Supelco, Inc., or equivalent
     to deactivate.)  Note:  Systems utilizing a jet separator for the
     GC effluent -are recommended since membrance separators may lose
     sensitivity for light  molecules and glass frit separators may
    'Inhibit the elutlon of polynuclear aromatics.  Any of these
     separators may be.used provided that it gives recognizable mass
     spectra (background corrected) and acceptable calibration points at
     the limit- of detection specified for each individual compound
     listed in Tables 4, 5, and 6.
4.3  A computer system must be interfaced to the mass • spectrometer to
     allow acquisition of continuous mass scans for the duration of the
     chromatographic program.  The computer system should also be
     equipped with mass storage devices^or saving all data from GC-MS
     runs.  There must be computer software available to allow searching
     any GC-MS mn for specific ions and plotting the intensity of the
     ions with respect to time or scan number.  The ability to integrate
     the area under any specific  ion plot peak is essential for
     quantification.
                          J/T

-------
    4.9  Continuous 11quid-11quid extractors - Teflon or glass  connnectlng
         joints and stopcocks, no lubrication.   (Hershberg-Wolf Extractor  -
         Ace Glass Co., Vine!and, N.J. P/N 6841-10 or equivalent).
5.  Reagents
    5.1  Sodium hydroxide - (ACS) 6N 1n distilled water.
    5.2  Sulfurtc add - (ACS) 6N 1n distilled water.
    5.3  Sodium sulfate - (ACS) granular anhydrous (rinsed with methylene
         chloride (20 ml/g) and conditioned at 400°C for 4 hrs.)
    5.4  Methylene chloride - Pesticide quality or equivalent.
    5.5  Stock standards - Obtain stock standard solutions at a
         concentration of 1.00 ug/ul. For example, dissolve 0.100 grans of
         assayed reference material 1n pesticide quality Isooctatte or other
         appropriate solvent and dilute to volume 1n a 100 ml ground glass
         stoppered volumetric flask.  The stock solution 1s transferred to
         15 ml Teflon lined screw cap vials, stored 1n a refrigerator, and
         checked frequently for signs of degradation or evaporation,
         especially just prior to preparing working standards from them.
         Protect PNA standards from light.
6.  Calibration Standards
    S'.l  Prepare calibration standards that contain the compounds of
         Interest, either singly or mixed together.  The standards should  be
         prepared at concentrations that will completely bracket the working
         range of the chromatographic system (tvro or more orders of
         magnitude are suggested).  If the limit of detection (Tables 4, 5,
         or 6) can be calculated as 20 ng injected, for example, prepare
         standards at 1 ug/ml, 10 ug/ral, 100 ug/ml, etc. so that injections

-------
     of 1-5 ul of the calibration standards will define the linearity  of
     the detector in the working range.
6.2  Assemble the necessary gas chromatographic apparatus and establish
     operating parameters equivalent to those indicated in Tables 4, 5,
     and 6.  By injecting calibration standards, establish the linear
     range of the analytical system and demonstrate that the analytical
     system meets the limits of detection requirements of Tables 4, 5,
     and 6.  If the sample gives peak areas above the working range,
     dilute and reanalyze.
6.3  The internal standard approach is acceptable for all of the
     semivolatile organics when internal standards meeting the following
     criteria can be selected:
     6.3.1  No interference with other components in the sample.
     6.3.2  Structural similarity to the compound to be determined.
     6.3.3  Elutes close to the compound to be  determined.
     6.3.4  The amount added approximates the concentration of the
            compound to be determined.
6.4  Internal Standard Method - The utilization of the internal standard
     method requires the periodic determination of response factors  (RF)
     which are defined 1n equation 1.
                    Eq. I  RF - (AsC1s)/(A1sCs)
     Where:
           A      is  the  integrated  area  or  peak  height of the
                 characteristic  ion for  the priority  pollutant standard.
           A.   is the  integrated area  or peak  height  of the
                 characteristic  ion for  the internal  standard.
                           ML.  -

-------
          C1s 1s the amount (ug) of the Internal standard.
          C_  1s the amount (ug) of the priority pollutant standard.
6.4  The relative response ratio for the priority pollutants should be
     known for at least two concentration values—20 ng Injected to
     approximate 10 ug/1 and 200 ng Injected to approximate the 100 ug/1
     level.  (Assuming 1 ml final volume and a 2 ul injection).  Those
     compounds that do not respond at either of these levels may be run
     at concentrations appropriate to their response.
     The response factor (RF) must be determined over all concentration
     ranges of standard (C$) which are being determined.  (Generally,
     the amount of Internal standard added to each extract is the same
     (20 ug) so that C1$ remains constant.)* This should be done by
     preparing a calibration curve where the response factor (RF) is
     plotted against the standard concentration (Cs).  Use a minimum
     of three concentrations over the range of interest.  Once this
     calibration curve has been determined, it should be verified daily
     by injecting at least one standard solution containing internal
     standard.  If significant drift has occurred, a new calibration
     curve must be constructed.  To quantify, add the internal standard
     to the concentrated sample extract no more than a few minutes
     before injecting into the SC/MS to minimize the possibility of
     losses due to evaporation, adsorption, or chemical reaction.
     Calculate the concentration by using the previous equations with
     the appropriate response factor taken from the calibration curve.
     (Ideally, the response factor will not ch'ange with concentration.)
     Suggested internal standrds and surrogate standards fall into two
                               23'7

-------
     groups.  A - Oeuterated compounds.  For Aromatics; benzene-dg,
     toluene-dg, and anthracene-dg.  For amines; pyridine-d-, and
     aniline-de.  For phenols; phenol-dg.  For nitroaromatics;
     nitrobenzene-dg,  B - Fluorinated compounds.  For aromatics;
     pentafluorobenzene, 1-fluoronaphthalene, 2-fluoronaphthalene,
     2-fluorobiphenyl, 2,2'-difluorobiphenyl, and 1,2,3,4,5-
     pentafluoroblphenyl.  For phenols; pentafluorophenol,
     2-perfluororaethyl phenol and 2-fluorophenol.  Compounds used as
     internal standards are not to be used as surrogate standards.  The
     internal standard must be different from the surrogate standards.
6.5  The external standard method can be used at the discretion of the
     analyst and is recommended when the criteria for use of the
     internal standard cannot be met.  Prepare a master calibration
     curve using a minimum of three standard solutions of each of the
     compounds that are to be measured.  Plot concentrations versus
     integrated areas or peak heights (selected characteristic ion for
     SC/MS).  One point on each curve should approach the limit of
     detection  (Tables 4, 5, and 6).  After the master set of instrument
     calibration curves have been established, they should be verified
     daily by injecting at least one standard solution.   If significant
     drift has occurred, a new calibration curve must be constructed.
Quality Control
7.1  Before  processing any samples, demonstrate through the analysis  of
     a method blank, that all glassware  and reagents are  interfersnce-
     free.   Each time a set  of samples  is extracted or there  is a change
     in reagents, a method blank should  be processed as a safeguard

-------
         against chronic laboratory contamination.
    7.2  Standard quality assurance practices should be used with this
         method.  Field replicates should be collected to determine the
         precision of the sampling technique.  Laboratory replicates should
         be analyzed to determine the precision of the analysis.  Fortified
         samples should be analyzed to determine the accuracy of the
         analysis.  Field blanks should be analyzed to check for
         contamination Introduced during sampling and transportation.
8.  Sample Collection, Preservation, and Handling
    8.1  Srab samples must be collected 1n glass containers.  Conventional
         sampling practices should be followed, except that the bottle must
         not be prerinsed with sample before collection.  Composite samples
         should be collected' In refrigerated glass containers in accordance
         with the requirements of the program.  Automatic sampling equipment
         must be free of tygon and other potential sources of contamination.
    8.2  The samples must be iced or refrigerated from the time of
         col-lection until extraction.  Chemical preservatives should not be
         used in the field unless more than 24 hours will elapse before
         delivery to the laboratory.  If the samples will not be extracted
         with 48 hours of collection, they must be preserved as follows:
         8.2.1  Add 35 mg of sodium thiosulfate per 1 ppm of free chlorine
                per liter of sample.
         3.2.2  Adjust the pH of the water sample to a pH of 7 to 10 using
                sodium hydroxide or sulfuric acid.  Record the volume of
                acid or base used.
    8.3  All samples must be extracted within 3 days and completely analyzad

-------
         within 7 days of collection.
9.  Sample Extraction (Base/Neutrals. Acids, and Pesticides)
    9.1  Mark the water meniscus on the side of the sample bottle for  later
         determination of sample volume.  Pour the entire sample into  a
         two-liter separatory funnel.  Adjust the pH of the sample with 6N
         NaOH to 11 or greater.  Use multirange pH paper for the
         measurements.  Proceed to Section 10 if continuous extraction is
         used.
    9.2_ Add 60 ml methylene chloride to the sample bottle, cap, and shake
         30 seconds to rinse the walls.  Transfer the solvent  into the
         separatory funnel, and extract the sample by shaking  the funnel for
         two minutes with periodic venting to release excess vapor
         pressure.  Allow the organic layer to separate front the water phase
         for a minimum of ten minutes.  If the emulsion interface between
         layers is more than one-third the size of the solvent layer,  the
         analyst must employ mechanical techniques to complete the phase
         separation.  The optimum technique depends upon the sample, but may
         include stirring, filtration of the emulsion through  glass wool, or
         centrifugation.  If the emulsion cannot be broken, that is,
         recovery of  less than 802 of the added solvent corrected for  the
         water solubility of methylene chloride, use continuous extraction.
         Collect the methylene chloride extract in a 250-ml Erlenmeyer flask.
    9.3  Add  a second 60-ml volume of methylene chloride to the sample
         bottle and complete the extraction procedure a second time,
         combining the extracts in the Erlenmeyer  flask.
    9.4  Perform  a third extraction  in the same manner.  Pour  the combined

-------
     extract through a drying column containing 3-4 Inches of anhydrous
     sodlun sulfate, and collect 1t in a 500 ml K-0 flask equipped with
     10 nl concentrator tube.  Rinse the Erlenmeyer with 20 to 40 ml of
     methylene chloride.  Pour this through the drying column.  Seal,
     label as base/neutral fraction, and proceed with the add
     extraction.
9.5  Add (Phenols) Extraction - Adjust the pH of the water, previously
     extracted for base-neutrals, with 6N H^SO^ to 2 or below.
     Serially extract with 60, 60, and 60 ml portions of
     d1stilled-1n-glass methylene chloride.  Collect and combine the
     extracts in a 250-
-------
     an apparent volume 1 ml, remove the K-0 apparatus and  allow  the
     solvent to drain for at least 10 minutes while cooling.  Remove the
     Snyder column and rinse the flask and Its lower joint  Into the
     concentrator tube with 1 to 2 oil of methylene chloride.  A 5-ml .
     syringe 1s recommended for this operation.
9.3  Add a clean boiling chip and attach a two-ball micro-Snyder  column
     to the concentrator tube 1n 9.7.  Prewet the column by adding about
     0.5 ml methylene chloride through the top.  Place the  K-0 apparatus
     on a warm water bath (60 to 65°C) so that the concentrator tube
     1s partially Immersed 1n the water.  Adjust the vertical position
     of the apparatus and the water temperature as required to complete
     the concentration in 5-10 minutes.  At the proper rate of
     •
     distillation the balls of the column actively chatter  but the
     chambers do not flood.  When the liquid reaches an apparent  volume
     of about 0.5 ml, remove the K-0 from the water bath  and  allow the
     solvent to drain and cool for at least 10 minutes.   Remove the
     raicro-Snyder column and rinse its lower joint into the concentrator
     tube witft approximately 0.2 ml of methylene chloride.  . Adjust the
     final volume to 1.0 ml.
9.9  Determine the original  sample volume by refilling the  sample bottle
     to the mark and transferring the liquid to a 1000 ml graduated
     cylinder.  Record the sample volume to the nearest 5 ml.
9.10 Pesticides.  Perform only if pesticides are found by GC-ECD.
     Carefully concentrate the 10 ml final  hexane extract (no dilutions)
     by "blowing down" to 0.5 ml using a clean  airstream.   Seal,  label
     as pesticide fraction.  Proceed to  section  11.

-------
10.  Emulsions/Continuous Extraction
    10.1 Place 100 to 150 ml of methylene chloride  1n the  extractor  and  ZOO-
         SCO ml methylene chloride 1n the distilling flask.
    10.2 Add the aqueous sample (pH 11 or greater)  to the  extractor.
         Extract for 24 hours.  Remove the distilling flask and pour the
         contents through a drying column containing 7 to  10 on of
         armhydrous sodium sulfate.  Collect the methylene chloride  in a 500
         ml K-0 evaporator flask equipped with a 10 ml concentrator  tube.
         Seal, label as the base/neutral fraction,  and concentrate as per
         section 9.7 to 9.10.
    10.3 Adjust the pH of the sample 1n the continuous extractor to  2 or
         below using 6H sulfurlc add. Charge a clean distilling flask with
         500 ml of methyl ene chloride..  Extract for 24 hours.  Remove the
         distilling flask and pour the contents through a'drying column
         containing 7 to 10 on of anhydrous sodium  sulfate.  Collect the
         methylene chloride layer on a K-fl evaporator flask equipped with a
         10 ml concentrator tube.  Label as the add fraction.  Concentrate
         as per section 9.7 to 9.10.
11. Gas Chromatoqrapny-Mass Spectrometry of Base/Neutrals  and Pesticide
    Fraction
    11.1 Tables 4, 5, and 6 summarize the recommended gas  chromatographic
         column materials and operating conditions  for the instrument.
         Included 1n this table are estimated retention times and
         sensitivities that can be achieved by this method.  Examples of the
         separations achieved by these columns are  shown  1n Figures  1
         through 7.  Calibrate the system daily with a minimum of one

-------
     injection of calibration standards.  Insure that the GC-MS system
     meets the criteria in reference 1 by injecting 50 ng of DFTPP
     through the GC inlet system.  If the extract must be stored
     overnight before analysis by GC/MS, transfer it with a disposable
     pipet, from the K-0 ampul to a solvent tight container.  Do not add
     the internal standard at this time.  The recommended container is a
     standard 2 ml serin vial equipped with a Teflon lined rubber septum
     and crimp cap.  These cap liners are inert and methylene chloride
     can be held without evaporation loss for months if caps liners are
     unpierced.  When the extracts are not being used for analysis,
     store them with unpierced septa in the dark at less than 4°C.
11.2 Internal Standard - Add 20 ug (10 ul of 2 ug/ul) of internal
     standard just prior to analysis.  Mix thoroughly.
11.3 Inject 2 to 5 ul of the sample extract.  The solvent-flush
     technique is preferred.  Smaller (1.0 ul) volumes can be injected
     if automatic devices are employed.  Record the volume injected to
     the nearest 0.05 ul, and the resulting peak size, in area units.
11.4 If the peak area exceeds the linear range of the system, dilute the
     extract and reanalyze.
11.5 At the beginning of each GC-MS run of a base-neutral fraction, the
     operator should demonstrate the ability to chromatograph 100 ng of
     benzldine.  Only after this is accomplished should the run be
     started.
11.5 At the beginning of each GC-MS run of the pesticide fraction, the
     operator should demonstrate the ability to chrcmatograph TOO ng of
     aldrin.  Only  after this should the run be started.

-------
12.  6as Chromatograohy/Hass Soectrometry of Add Fraction
     12.1  Before at acid extract Is analyzed by 6C-MS, the operator should
           demonstrate the ability to detect 50 ng of pentachlorophenol.
     12.2  After acceptable Instrumental response 1s demonstrated, proceed
           with the may sis as described for the base/neutral fraction^
           (11.1 to 11.4).
     12.3  Table 5 lists the retention times of the 11 aoid extractables.
           Chroraatography of nltrophenols 1s. poor.  The limits of detection
           given refer to the amounts required to obtain MS confirmation by
           the methods described below.
13.  Mass Spectrometry
     13.1  The complete background corrected mass spectrum 1s compared to a
           reference mass spectrum, either from an authentic sample or a
           library spectrum, to provide qualitative Identification.  All of
           the following criteria must be met.
           13.1.1  The experimental mass spectrum must contain all ions
                   present above 1% relative abundance in the mass spectrum
                   of the reference with the general agreement of * 2QX of
                   the relative abundance 1n the reference mass spectrum.
           13.1.2  Ions present in the experimental mass spectrum that are
                   not present in the reference mass spectrum must not
                   exceed 10% of the total ion abundance in the experimental
                   mass spectrum.  These contaminant ions should be readily
                   accounted for by common background Ions.  If the
                   experimental mass spectrum is highly contaminated,
                   identification should not be accepted.

-------
           13.1.3  The retention time at the experimental mass spectrum
                   should be within ^ 5 spectrum numbers or * 30 seconds
                   (whichever 1s greater) of the retention time of the
                   authentic compound.
           13.1.4  Structural isomers that have very similar mass spectra
                   (phenanthrene, anthracene; faenzo(a}pyrene,
                   benzo(e)pyrene, etc.) can be explicitly identified if the
                   resolution between the isomers in a standard mix is
                   acceptable.  Acceptable resolution is achieved if the
                                                              2.5% f
                   valley height between isomers is less than -freW- the sum
                   of the two peak heights.  Otherwise, structural isomers
                   are identified as isomeric pairs.
     13.2  In samples that contain an inordinate number of interferences the
           chemical ionization (CI) mass spectrum may make identification
           easier.  In Tables 7 and 8 characteristic CI ions for most of the
           compounds are given.  The use of chemical ionization MS to
           support El is encouraged but not required.
14.  Calculations (Base/Neutrals and'Acids)
     14.1  When a compound has been identified, the quantification of that
           compound will be based on the integrated area from the specific
           ion plot of the first listed characteristic ion in Tables 7, 8
           and 9.  If the sample produces an interference for the first
           listed ion, use a secondary ion to quantify.  Quantification will
           be done by the external or internal standard method.
     14.2  Internal Standard - By adding a constant known amount of  internal
           standard (C-  in ug) to every sample extract, the concentration

                                •ML-  2H ±

-------
      of priority pollutant (CQ) 1n ug/1 in the sample is calculated
      using equation 2.
                r-  ,    ,   .      
                 ^        0      (A ) (RFj (Voj
      Where:                       s
            V  1s the volume of the original sample in liters.
14.3  External Standard - The concentration of the unknown can be
      calculated from the slope and intercept of the calibration
      curve.  The unknown concentration can be determined using
      equation 3.
                                                   (A)(VJ
         Bj.  3     Mlcrograrcs/liter * ng/ml »     ,\> Wy i
      Where:
           A » mass of compound from calibration curve (ng).
           V, » volume of extract injected (ul)
           Vt » volume of total extract (ul)
           V  « volume of water extracted (ml)
14.4  Report all results to two significant figures.  Report results in
      raicrograms per liter (Base/Neutrals and Adds) without correction
      for recovery data.  Report qualitative identification only for
      pesticides (quantitative data is available from GC-SC).  When
      duplicate and spiked samples are analyzed, all data obtained
      should be reported.  Relative response ratios and response
      factors from MS analysis should be included when reporting data.
      Include identification of any components found in the blanks.
14.5  In order to minimize unnecessary GC-MS analysis of method blanks
      and field blanks, the field blank may be run on a FID-GC equipped

                          JIT- o?^

-------
           with the appropriate SP-2250 or SP-1240 OA columns.  If no peaks
           are seen of Intensities equal to or greater than the internal
           standard, then it is not necessary to do a GC-MS analysis of the
           blank.  If such peaks are seen, then the field blank must be sent
           for full priority pollutant analysis.
15.  Accuracy and Precision
     15.1  Data not available at this time.
16.  Quality Assurance
     16.1  GC-MS system performance evaluation is required each day the
           system is used for samples or reagent blanks.  A sample of 50 ng
           of decafluorotriphenylphosphine (PCR, Gainesville, Florida) is
           injected into the system, and the mass spectrum is acquired and
           the background corrected spectra is plotted.  Criteria
           established in Reference 1 must be met.  The analyst should also
           •demonstrate that the analytical conditions employed result in
           sharp total ion current peaks for 100 ng of benzidine on the
           SP-2250 column when this column is used for Base/Neutrals, 100 ng
           of aldrin on the SP-2250 column when used for pesticides, and 50
           ng of pentachlorophenol on the  SP 1240-DA column when it is
           used.  All plots from the performance evaluation must be retained
           as proof of valid performance.  See Reference 3.

-------
17.  References

1.   Method 610, Polynuclear Aromatic Hydrocarbons, U.S. EPA, EMSL,
     Cincinnati, Ohio, 45268, 1979.
2.   "Reference Compound to Calibrate Ion Abundance Measurement 1n Gas
     Chromatography—Mass Spectrometry Systems," J. W. Elchelberger, L. E.
     Harris and W. L. Budde, Anal. Chen. 47, 995-1000 (1975).
3.   "Quality Assurance and Quality Control Procedures for Screening and
     Verification of Industrial Effluents for Priority Pollutants," U.S.
     EPA, EMSL, Cincinnati, Ohio, 45268, 1979.

-------
                           TABLE 1
                  Base-Neutral Extractables
 Compound

Acenaphthene
Acenaphthylene
Anthracene
Benzo(a)anthracene
Benzo(b)f!uoranthene
Benzo(k)f1uoranthene
8enzo(a)pyrene
8enzo(gni)perylene
Benzldine
B1s(2-chloroethyl)ether
B1s(2-ch1oroethoxy)methane
B1s(2-«thy1hexyl)phthalate
81s(2-chloroisopropyl)ether
4-8romcpheny1 phenyl ether
Butyl benzyl phthalate
2-£h1oronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
D1benzo(a,h)anthracene
D1-n-butylpnthalate
1,3-Qichlorobenzene
1,4-Oichlorobenzene
1,2-01ch1orob enzene
3,3'-01chlorobenzidine
Dlethylphthalate
Dimethylphthalate
2,4-01n1trotoluene
2,6-01nitrotoluene
Dloctylphthalate
1,2-01phenylhydrazi ne
Fluoranthene
Fluorene
Hexachlorobenzene
Hexach1orobutadiene
Hexach1oroeth ane
Hexachlorocyclopentadiene
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-ni trosodimethy1 amine
N-nitrosodi-n-propylamine
N-n1trosod1phenylamine
Phenanthrene
Pyrene
2,3,7,8-Tetrachlorodibenzo-p-dioxin
1,2,4-Trichlorobenzene
STDRET No.

  34205
  34200
  34220
  34526
  34230
  34242
  34247
  34521
  39120
  34273
  34278
  39100
  34283
  34636
  34292
  34581
  34641
  34320
  34556
  39110
  34566
  34571
  34536
  34631
  34336
  34341
  34611
  34625
  34596
  34346
  34376
  34381
  39700
  34391
  34396
  34336
  34403
  34408
  34696
  34447
  34438
  34428
  34433
  34461
  34469
  34675
  34551
                      IT-

-------
                         TABLE 2
                    Add Extractables
    Compound                             Storet No.

4-Chloro-3Hnethylph«nol                     34452
2-Chlorophenol                              34536
2,4-01ch1orophenol                          34601
2,4-Olmethylphenol                          34606
2,4-01n1trophenol                           34616
2-Methy1-4,6-/

-------
       Table 3

Pesticide Extractables
Aldrln
a-BHC
b-BHC
d-BHC
g-BHC
Chlordane
4, 4 '-ODD
4, 4 '-ODE
4,4' -DOT
01eldr1n
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrln
EndHn Aldehyde
Heptachlor
Heptachlor Epoxide
Toxaphene
PC3-1016
PCS- 1221
PCS- 1232
PCB-1242
PCS- 1248
PCB-1254
PCS- 1260
                        39330
                        39337
                        39338
                        39259
                        39340
                        39350
                        39310
                        39320
                        39300
                        39380
                        34361
                        34356
                        34351
                        39390
                        34366
                        39410
                        39420
                        39400
                        34671
                        39488
                        39492
                        39496
                        39500
                        39504
                        39508

-------

                oo o oo oo oo ooooo o o o oo oo ooooo o o o o
 vt
  o» » ce « r*» u) to co •—

             »^ cJ cj tn ut p* r^ oo flo o» oJ o»
                                                                             •— u> o » O
                      ••  ••
                      41  41

                     5-s

                                                   2
                                                   41
              •a


           iw
           3^5
                  «    I
        •3   *    |    5


        1   1   i    *
v 4>   o> «
                                                                             «s c
                                                                                             a
                                                                                             •w
                                                                                             «

                                                                                             *«
                                                        e   -co
                                                        4) 4> 44 44
                                                        r— C.C O
                                                      C *•»•—.£ Cb4)  >i «
                                                      41 C >t^4 •— ^  C C
                                                     js s. e j=-o ^  aj iCci.c  a<~ o»— a -i->  u 3
    I  O  ^  I JS» O CL ^L^C *^" 41 O **^ "^ ^^ W ^S S C  *O ^3
   P ^5 ^5 CM U ^* fO 'Q 44 C3 ^ ^^ Cd CA ^S 44  u Q (O  ^  i

 X CM O  ^L ^) X ^J 41 41 B ^O 3 *,J ^p CM 41 C  X C3 42 ^4  I
 0}  • tn  rtj •»- 
-------
*-•   o o O O O O O O o o o o in in in
 O)   r* r* v+*+r+r* r+f+v* r+i-l 1*4 f*t &l fit
 3
                                                                                          VI
                                                                                                •
                                                                                                    (IS
                                                                                                    s  o
 Ol
•M
 U

•2?   O OO O O O O O O OO O O O O
                                                                                      1-4 It)
                                                                                      >
                                                                                      O
                                                                                      O
                                                                                                      •o  o
•o
 »«-» > M -
 31-CJ->U1!-C(T
                           I  O O O
                             Ciki fci i^j
                             r^i I^t fl
       c-wviuctn   i  c c c •


u.o.cacacQ«_iearoocacaca
                                                                                          OJ
                                                                                       e o
                                                                                       O I/I
                                                                                       01
                                                                                      *i
                                                                                       )Q
                                                                                       o
                                                                                      O  O
                                                                                      in  u
                                                                                      CM  a.
                                                                                      CM
                                                                                       I  OJ
                                                                                      a.  i.
                                                                                      «/1  3
                                                                                             vi
                                                                                             01
                                                                                                    
                                                                                      i. *J  O
                                                                                          u +•»
                                                                                      01  -^-o
                                                                                      ^-  e      "O
                                                                                      w?     C
                                                                                           .  «
                                                                                      F—  VI
                                                                                      « -i_> -n
                                                                                      u  e  
                                                                                      cox
                                                                                      « f-  
                                                                                      — ^a      0)
                                                                                                    OJ. — .-<-»

                                                                                          01 •
 3tl~ S


 01  .en
.* c

 ^go
 CL    <*-

""* 01 O
O 0.0
                                                                                       I SCM

                                                                                         a •»-»
                                                                                      CM
                                                                                                   o
                                                                                                                       OJ
                                 •a
                                 cu
                                 u
                                 o
                                 .«»
                                 0)

-------
                TABLE 5
Gas Chromatography of Add Extractables
               Retention Time*
                   (M1n)
                                                     Limit of Detection*
  Compound                            (M1n)           nq Injected        uq/1

                                                                         25
                                                                         25
                                                                         25
                                                                         25
                                                                         25
                                                                         25
                                                                         25
                                                                        250
                                                                        250
                                                                         25
                                                                         25


*6 foot glass column (1/4 1n. 00 x 2 mm 10} packed with 1% SP-1240 DA coated
 on 100/120 mesh Supelcoport.  Carrier gas:  helium at 30 ml per m1n.
 Temperature program:  2 m1n. Isothermal at 70°, then 8° per m1n to
 200QC.

*Th1s Is a minimum level at which the entire analytical system must give
 recognizable mass spectra (background corrected) and acceptable'calibration
 points.   NanVans Injected 1s based on a 2 ul Injection of a one liter
 sample that has been extracted and concentrated to 1.0 ml.
2-Chlorophenol
2-N1tropheno1
Phenol
2, 4-Q1methy 1 phenol
2,4-01chlorophenol
2, 4, 6-Trf chlorophenol
4-Chl oro-3-methylphenol
2,4-01 nltrophenol
2-Methyl -4,6-d1n1trophenol
Pent ach 1 orophenol
4-N1trophenol
5.9
6.4
3.0
9.4
9.3
11.3
13.2
15.9
16.2
17.5
20.3
50
50
50
50
50
50
50
500
500
50
50

-------
                                    Table 6

                  Gas Chromatography of Pesticide Extractables


                           Retention                Limit of Detection*
Compound                   Time (min)*              ng  injected     ug/1

a-bhc                        21.1                       40                10
g-bhc                        22.4                       40                10
b-bhc                        23.4                       40                10
Heptachlor                   23.4                       40                10
d-bhc                        23.7                       40                10
Aldrin                       24.0                       40                10
Heptachlor epoxide           25.6                       40                10
Endosulfan I                 26.4                       40                10
Oieldrin                     27.2                       40                10
4,4'-ODE                     27.2                       40                10
Endrin                       27.9                       40                10
Endosulfan II                28.5                       40                10
4,4'-000                     28.6                       40                10
4,4'-ODT                     29.3                       40                10
Endosulfan sulfate           29.3                       40                TO
Cnlordane                    19 to 30
Toxaphene                    25 to 34
PCB-1242                     20 to 32
PC3-1254                     23 to 32

*6 foot glass column (% in OF X 2 mm ID) packed with  3% SP-2250  coated on
 100/120 mesh. Supelcoport.  Carrier gas:  helium at 30  ml per min.
 Temperature Program:  Isothermal for 4 minutes at 50°C, then 8° per
 minute to 270°.  Hold at 270°C for 5 minutes.

#Tn1s is a minimum  level at which the entire  analytical  system must given
 recognizable background corrected mass spectra.  Manograms  injected is  based
 on a 2 ul injection of a one-liter sample that has been extracted  and concen-
 trated to a volume of 0.5 ml.
                                  :nr-

-------
   S
   M


   41
   •o>
   U

   w
ui
   Ml
CD ^
   •
   o


           O O (^ C) O I
   N      U) UI O O UJ i         	  - 	 		  ___          ___    _  	
           o4 o4 CM 04 o4 o4    o4o4CMCMo4o4CMCMo4o4o4CMo4   CMCMCMo4*CMCMCMCMCM

   Q C

           v v O O ^r en    so us CM co us O f o» us us vo 04 vo   o« 04 CM n» ca ui P»  r^ o
           04 04 CM 04 04 04    o4 o4 CM 04 04 04 CM 04 04 o4 04 CM 04   CM CM CM o4 CM CM 04  04 CM



           S^O VO' Crt    ^O ^^    crt *^* ^^ o4 C^l ^fi U3 ^n CM *^P o4 f^ IO   ^n U9 ^^ €ft ^y Gft CO  CO ' _
           ^r ^r en en ^r f*1*    en CM CM C9 CM vo en vo u) uiuicovo   co eo f^ vo 03 ^ ^»  r^ 4t>
^5      040404U304      ^,04 CM 04 04   CM 04 04 04 04 o4 O4   o4 04 04 04 CM CM 04  04 04








4)
4->
u •«
«o u          	
s. re«+4   o4o»cncno4i
a ex-i •

gofj!
           00 CO *^ f^ 09 ^^ CM    ^^ ^^ ^^ O) U3 UB ^^ **4 fO ^* OH U) ^O ^^ ^^ ^H CO ^J ^3 O^  ^ft *^

   2*t    '~^'CN*    ^      Cr»o4CMo4o4   CM o4 04 04 o4    04CM       o4 04 o4 CM o4  ,


   U'
   4>

                      04   04      CMo4o4~CMo4o4o404o404CMo4    l^«t^CMCMo4<
   U
   l«
   s.
   •M
   X
   4>
   UI
   "S
                                   1
                                   4J
                                        4) S          4)
*•       4)  4J

c c C .e c j: 2     -a "3
4J 4) re i^ V VJ
.a .a .c  ,
JI.C O JS -C ^
u u <— o u -w

O O U CM O S
 I   I  <9»^ I  •*—«
m ^> x ut CM »
                                                       X
                                                       O
                                 
                                                               C
                                                               O)
                                                                             41
                                                                             •H
                                                                             «
                                                                             4)
                                                                                             
VI -o>
                        I «— H- 
                              4)
                              C
                              0)

                              5
                              a
                              -M
                              a
                                                                        c e
                                                                       •r- ,'o^1
 ai a
J= s. c
 C.-W » -
 S C "3  >l
 a «s t.  -w

CO 01 *J  ^J
 I  JS C  -—
^r a.<  a

-------
   e . r*» r»»
                                           CM CM CM CM CM CM
                                           CM CM CM VO CO IO
                                           in in in r>. r«» r*.
                                           CM CM CM CM CM CM
                                                               CM
en
CO
01
4^
u
   u
   i«— >) M -  U N N M
            O 4) M >i«— >) M -  U N N M , 4) 3 — JS  4J   •«>- 4J 4> 4) G
                        41

                        19



                     41 4-1       4)
                    4J j=       e
                     «  a.      —    4> 4>    i.
                    •—          -o    c c    >»
                     lQ<»~k   41 i—    4) 4)    O,
                    £.<—    C M 4J.C J=   —-.
                    4^>,   4J C -M -M J-l   T3
                    J5X    U4>«CC     <•
                     a. 4i     -        " ~
                                                                                     4)


                                                                                  4)1
                                                                                  e t)
                                                                                  oi <—
                                                                                  IN >l

                                                                                  4) O)


                                                                                  If
                                                                                  
-------
                                ,•« v>4 «•*** 1-4 CM *-« ^«
 in

 a
 01




 41
I t,


 g

\ 4>
01
•M

8$

31*
   a

   j^
   u
   01-
                     *^t^   f^   €NJ i-* i««    PU «•«
             CM

 U

 10
              o



              u
                                  o


                            i    I
                          «—4i    a.
                          o .c    a

                     — §*   i
                        §0 .C ,— — ,_.—   ^-,
                        «s a. >» a e  o   o
                      ~o  »JS Q U

             >— jj  a a o >o •-• o *I  >o4-> w
             J= —  C  I  I   •£  I  OJ -ta> — .C
             o Z    1  C
             CMCMQ.CMCMCM^>CMCM^^><

-------
                                Table 9

                     Pesticides Characteristic  Ions
                                        Characteristic Ions
                                           Electron  Impact
Compound                                      fritli  Inti]
a-bhc                                     183      131      109
g-bhc                                     183      181      109
b-bhc                                     181      183      109
heptadilor                                100      272      274
d-bhc                                     133      109      181
aldrln                                    66      263      220
heptachlor ep oxide                        353     . 355      351
endosulfan I                              201      283      273
dleldrln                                  79      263      279
4,4'-QOE                                  246      248      176
4,4'-000                                  235      165
endrln                                    81      263       82
endosulfan -II                             201      283      278
4,4'-QOT                                  235      237      165
endosulfan sulfate                        272      387      422
chlordane*                                373      375      377
toxaphene**                               231      233      235
PCB-1242**                                224      260      294
PCB-1254**                                294      330      362

*  Characteristic of  alpha  and gamma  forms  of chlordane.
**     .                               ,> theKcompounds are mixtures of
   various Isomers.

-------
        o
                                                                 CM
                  3N3>IH1NVN3H
o



o
                                                                         3
                                                                         a

-------
        10N3HdOHilN-fr
   10N3HdOH01HOViN3d
        10N3HdOailNia-tr'2
10N3HdOa01HO IHi-9' fr 'Z
10N3HdlAH13Wld-fr'Z
    = K  10N3HdOa01HO-2
                                                      e
                                                      o

                                                      u
                                                      
-------
    aaa-.fr> « n


aaa-.fr>
iaa-.fr>
       =s

           Ninons
           •^••M
           MM^BM

 I NVdinsoaivs
                                                             u
                                                             
-------
COLUMN: 3% SP-2250 ON SUPaCOPORT
PROGRAM: 50*C. 4 MIN. 8*PER WIN TO 270"C
DETECTOR: MASS SPECTROMETER
                                  •PEAKS GIVING THE THREE
                                   CHARACTERISTIC IONS

20
25
                                         30
                RETENTION TIME-MINUTES

Figure 4.  Gas chromatogram of chiordane

-------
   COLUMN: 3% SP-2250 ON SUP&COPORT
   PROGRAM: 50*C, 4 WIN. 8*PER MIN TO 27Q*C
   DETECTOR: MASS SPECTROMETER
           25             30            35
                 RETENTION TIME-MINUTES

Rgure 5. Gas chromatogram of toxaphene

-------
COLUMN: 3% SP-2250 ON SUPELCOPORT
PROGRAM: 50«C. 4 WIN. 8°PER WIN TO 270"C
DETECTOR: MASS SPECTROMETER
                                    A m/ft 224 PRESENT
                                    8 m/e 260 PRESENT
                                    C m/e 294 PRESENT
           20
   25            30
RETENTION TIME-MINUTES
35
 Figure 6. Gas chromatogram of Arochlor 1248

-------
  COLUMN: 3% SP-2250 ON SUPELCOPERT
  PROGRAM: 50*C, 4 WIN, 8* PER MIN TO 270aC
   DETECTOR: MASS SPECTROMETER
   A n/e 294 PRESENT
   B m/e 330 PRESENT
   C m/e 362 PRESENT
            20            25            30
                 RETENTION TIME-MINUTES

Rgure 7. Gas chromatogram of Arochlor 1254

-------
 QUALITY ASSURANCE AND QUALITY CONTROL PROCEDURES FOR
SCREENING AND VERIFICATION OF INDUSTRIAL EFFLUENTS FOR
                 PRIORITY POLLUTANTS
        U. S. ENVIRONMENTAL PROTECTION AGENCY
   ENVIRONMENTAL MONITORING AND SUPPORT LABORATORY
               CINCINNATI, OHIO  45268
                      June 1979

-------
                                 INTRODUCTION

    "The Sampling and Analysis Procedures for Screening of Industrial
Effluents for Priority Pollutants," (revised April, 1977) was designed to
provide qualitative/seraiquantitative determination of the priority
pollutants.  In order to apply these procedures to quantitative analysis,
adequate quality assurance practices must be incorporated into the
procedures.  This document provides modifications to enable  implementation
of a quality assurance program that will define the accuracy and precision
of the procedures for specific parameters.

    In considering modifications to be made, the following assumptions were
made:

    1.  The program must be well documented.  It must be integrated  into
        field and laboratory procedures.  It must include field and
        laboratory blanks, equipment calibration checks, replicate and
        spiked sample analyses, the analysis of check standards and
        performance evaluation standards.

    2.  The differences in industrial subcategories require method
        validation and the establishment of control parameters for each
        subcategory.

    3.  Verification Phase analyses will seldom, if ever, require analysis
        of all 114 organic priority pollutants.  Thus, the quality control
        program-will focus on a Smaller selected number of compounds based
        on the results of screening analyses and other data.

    4.  For purposes of the Effluent Guidelines Verfication  Phase studies,
        the quality control programs provided, herein, must  be applied to
        the sampling sources most  frequently involved in rule making,  i.e.,
        supply water, influent to  treatment, and effluent from treatment.
        Primary emphasis should be placed on the effluent from treatment
        since this is where the control limits will be set,

    5.  Two distinct levels of quality control are required:   initial  method
        validation and continuing  performance evaluation.  The initial
        quality control is to demonstrate the method precision and  accuracy
        and thus  validate the method on a common wasta stream  (industrial
        subcategory) where waste characterization was the basis"for
        establishing the subcatagory.

    5.  The  initial control limits will be  established for each  subcategory
        on the  effluent from treatment prior to initiation of  the routine
        sampling  and analysis program for that subcatagory.  The  control
         limits  for supply water and influent to treatment watar will be
        established over a period  of time using data derived from the
        ongoing continuing quality control  program.

    7.  As more  quality  control data  is obtained  for each wasta  type within
        each  subcatagory, the precision and  accuracy of  the  method  will  be
        updated  to provide better  defined control  limits.

                             inr-  j

-------
                                   FOREWORD

    These methods for sampling and analysis of  Industrial wastes  have been
prepared by the staff of the Environmental Monitoring  and Support
Laboratory, at the request of the Effluent Guidelines  Division, Office of
Water and Hazardous Wastes, with the cooperation of the Environmental
Research Laboratory, Athens, Georgia.  The Quality Assurance/Quality  Control
(QA/QC) procedures were prepared by the staff of the Environmental
Monitoring and Support Laboratory with the cooperation of the Effluent
Guidelines Division, Office of Water and Hazardous Wastes, and the
Environmental Research Laboratory, Athens, Georgia, And several EPA Regional
Laboratories.

    The procedures represent the current state-of-the-art and are applicable
to the Effluent Guidelines'Survey of prirority pollutants.  This  survey is
divided into two phases; the Screening Phase and the Verification Phase.

    The objectives of the Screening Phase are:

    1. To provide a deternrination of the presence or absence of a particular
       priority pollutant.

    2. To provide a quantitative estimate of the concentration of priority
       pollutants present.

    3. To focus the verification survey to specific priority pollutants.

    4. To provide direction to other best available technology (BAT)
       regulatory strategies.

    The objectives of the Verification Phase are:

    1. To provide the confirmation of the presence or  absence of  the
       priority pollutants Identified during the screening phase, necessary
       for consideration in rule making.

    2. To provide qualitative pollutant data with a known precision and
       accuracy.

    3. To provide pollutant data for use in rule making.

    $. To serve as a basis for technology selection.

-------
                                  DISCLAIMER

    The mention of trade names or commercial products 1n this document 1s
for Illustration purposes, and does not constitute endorsement or
recomnendatlon for use by the U. S. Environmental Protection Agency.

-------
         Quality Assurance and Quality Control Procedures
      for Screening Analysis for Organic Priority Pollutants
110 Scope
    This procedure 1s provided for use by laboratories
    performing screening analyses using the GC/MS methods 624
    and 625.  Screening analysis provides a determination of the
    presence or absence of a specific compound and a
    searfquantitative estimate of the concentration of a specific
    compound.  Quality Assurance (QA) 1s the total program for
    assuring the reliability of the monitoring data.  Quality
    control (QC) 1s the routtne application of procedures for
    controlling the measurement process.
120 Routine Quality Assurance and Quality Control
    121  Method Blank - The method blank 1s defined as an
         appropriate volume of "organic-free" water which has
         been processed exactly as the sample (Including
         glassware, reagents, solvents, etc.).  Reagents or
         solvents having background levels that Interfere with
         the compounds to be determined must be purified and
         shown to be acceptable or replaced with some that are
         acceptable prior to proceeding with analyses.  Problems
         encountered and corrective actions taken shall be
         documented and reported for the record.
         121.1  For the extractafale fractions (Base/Neutral,
                Add, and Pesticide) the method blank requires

-------
           extraction of  1-liter  "organic-free" water.   A
           method blank must be extracted  for  each  set  of
           field samples  extracted  at  a  given  time  (at
           least one method blank per  20 field samples
           analyzed) and  whenever a new  source of reagent
           or  solvent 1s  Introduced Into the analytical
           scheme.  The method blank can be screened by
           GC-FID.  Analysis by GC-MS  is required only  1f
           GC-FID analysis of the field  blank  gives peaks
           larger than the Internal standard peak.
     121.2 For the  volatile fraction,  5  ml of  "organic-
           free" water should be  analyzed  by the  purge  and
           trap methodology only  1f positive  Interferences
           are noted during the  analysis of a  field blank.
           If  positive  Interference stm  occurs, repeat
           the method blank analysis.   If  interference
           persists, dismantle the  system, thoroughly clean
           all parts that contact the  sample,  purge gas,
           and carrier  gas.  Replace or repack the  sorbent
           trap and change purge  and carrier gas.
122  Field Blank
     The field  blank is  defined  as an appropriate volume of
                                       sent
     "organic-free"  water  which has  been^to the sampling
     site and back  to  the  analytical laboratory in a
     container  and  bottle  identical  to  the  type used to
     collect  the samples.   Field  blanks and samples  must be

-------
     shipped 1n separate containers.  When received In the
     lab, the field blank 1s dosed, extracted and
     concentrated as If It were an actual sample.
     122.1  For the extractable fractions (Base/Neutral,
            Add, and Pesticide),  the field blank may be
            screened by GC-FID or  SC-EC (pesticides only).
            Full  SC-MS analysis 1s to be performed 1f the
            screening analysis gives any peaks larger than
           .the Internal standard  peak.
     122.2  For the volatile fraction, analyze a 5-ml field
            blank after each sample analysis.  Follow the
            guidelines 1n 121.2 1f positive Interferences
            are noted during the analysis of a field blank.
123  GC-MS Calibration Chedc
     123.1  For the Base/Neutral-Acid or Pesticide fraction,
            dally analyze decaf1uorotr1pheny1phosph1ne
            (OFTPP) by Injection Into the GC Inlet.  See
            Reference 1.  (DFTPP must also be run after any
            mass spectrometer tuning 1s done).  This
            calibration check may  be done fn conjunction
            with Section 124.  The requirement Is that 50
            nanograms of DFTPP Injected must meet the
            specifications prescribed.  Table 1 lists the
            OFTPP key Ions and 1on abundance criteria.
     122.2  For the volatile (purgeable) fraction, analyze
            4-bromo-fluorcbenzene  (BFB) daily fay direct

-------
            injection Into the GC/MS.  The requirement is
            that 20 nanograms of PFB8 injected must meet the
            prescribed specifications.  Table 2 lists the
            BFB key ions and ion abundance criteria.
124  6C Performance Check
     124.1  SP-2250 column - At the beginning of each day
            that base/neutral or pesticide analyses are to
            be performed, inject 100 nanograms of benzidine
            (base neutral analyses) or aldrin (pesticide
            -analyses) either separately or as part of a
            standard mixture that may also contain 50 ng of
            DFTPP.  Performance for base/neutrals is
            acceptable if the tailing factor for benzidine
            is less than 3.  Performance for pesticides is
            acceptable if the tailing factor for aldrin is
            less than 2.  Calculation of the tailing factor
            is given in Reference 2 and described in Figure
            1.  Retain all p-lots of performance evaluation
            as evidence of valid performance.
     124.2  SP-1240 OA Column - At the beginning of each day
            that acid fraction analyses are to be performed,
            inject 100 nanograms of pentachlorophenol either
            separately or as part of a standard mixture that
            may also contain DFTPP.  Performance is
            acceptable if the tailing factor is less than
            5,  Calculation  of the tailing factor is given
            in Reference 2 and described in Figure  1.
            Retain all plots of performance evaluation  as
            evidence of valid performance.

-------
         124.3   Carbopak C with Carbowax T500 or Carbopak B with
                SP-1000 column.  At the beginning of each day
                that purgeables analyses are to be performed,
                Inject 20 nanograms of BFB either separately or
                as part of a standard mixture.  Performance 1s
                acceptable If the tailing factor 1s less- than
                2.  Calculation of the tailing factor 1s given
                1n Reference 2 and described in Figure 1.
                Retain all plots of performance evaluation as
                evidence of valid performance.
    125  Performance Evaluation
         Once each 20 working days; dose, extract, and analyze a
         performance evaluation sample (available from
         EMSL-C1nc1nnat1).  Retain the results as evidence of
         valid performance.
    126  If the final extract 1s to be retained, or shipped to
         another location, follow the appropriate guidelines
         given in Appendix A.
130 Reporting of Data
    13T  All concentrations should be reported in ranges, i.e.
         less than 10 ug/1, 10 to 100 ug/1, or greater than 100
         ug/1.
    132  Report the mean and standard deviation of the relative
         response ratios of the standards (10 and 100 ug/1) from
         the MS analysis with the range values in 131.
    133  Report all quality control data with the analytical
         results.  This includes, but is not limited to, the
         recovery of surrogate standard spikes (Method 624), the
         precision for the analysis of surrogate standard spikes
                         or-

-------
         (Method 624), the results from the analysis of
         performance evaluation samples, and Identification of
         components found 1n the blanks.
    134  If the GC/MS data 1s to be archived on magnetic tape,
         follow the guidelines given 1n Appendix 3.
200 Sampling Procedures for Screening Phase
   •The Initial characterization (screening) of the varied
    Industrial discharges covered by this program will be made
    on an analysis of a. composite effluent sample.  Any scheme
    for collecting a composite sample is, in effect, a method
    for mechanically integrating to obtain average
    characteristics of a discharge.  During the screening phase
    the sample composite can be used to determine the average
    characteristics which would be representative of that
    discharge.  Simple composite samples are those that are made
    up of a series of aliquots of constant volume collected at
    regular time intervals in a single container.  Some
    situations may require flow or time proportional sampling;
    this determination will be made by the individual project
    officer after considering his specific industrial category.
    The determination of compositing period 24, 48, or 72 hours
    will be made on a case-by-case basis.  The duration of
    compositing will depend on the type of sample being
    collected, the type of facility being sampled, and the time
    varying characteristics of the discharge.  The rate of
    change of flow and other characteristics of the discharge

-------
and the accuracy required will also Influence the
determination of the compositing period.  For example,
longer compositing periods would be warranted when less
stable unit process operations are being sampled.
All samples and blanks must be carefully Identified using
water proof labels and water proof Ink.  Include the
following Information on the label:  sample number, date and
hour of sampling, complete Information as to source and
sampling point, preservative added, if any, and name of
person collecting the sample (Include address and/or phone
number).
201  Extractable Qrqanlcs (Base/Neutrals -' Acids and
     Pesticides)
     201.1  Collect a representative composite sample.  The
            maximum time Interval between aliquot samples
            shall be no longer than 30 minutes.  The minimum
            aliquot size shall be 100 ml.  The sample must
            be collected with an automatic sampler using the
            equipment and methods outlined below.  Minimum
            composite volume must be 2% gallons.
     201.2  Automatic Sample Collection - A peristaltic pump
            automatic sampler with timer and a single glass
            compositing jug is required.  The 2h - 3 gallon
            compositing battle must be glass and cleaned as
            outlined below.  New unused tubing must be used
            for the sampling line and

-------
       for the pump  for each Individual  outfall or
       sample location.  Vacuum type automatic samplers
       may be used provided that the sample chambers
       are glass and that they are cleaned after every
       use as outlined for glass composite containers.
       Place the sampler or composite container in an
       Insulated chest and ice.  Maintain the sample at
       4°C during the compositing procedure.  At the
       completion of the compositing period, seal the
       container with a Teflon lined cap.  Place the
       container in  an insulated shipping container,
       ice, and seal, then ship to the analytical
       laboratory.  Maintain at 4°C during transport
       and storage prior to analysis.
201.3  When sampling raw untreated industrial
       discharges which are generally high in suspended
       solids, it is imperative that adequate sample
       flow rate be maintained throughout the sample
       train  in order  to effectively transport the
       solids.  In horizontal runs, the velocity must
       exceed the scour velocity, while in vertical
       runs the settling or the fall velocity must be
       exceeded several times to  assure adequate
       transport of solids  in the flow.  The equipment
       used  in  sampling raw discharges then must have  a

-------
            minimum Intake velocity of 60 on per second  (2
            fps).  In the sampling of treated effluents,
            Just about any commercially available automatic
            liquid sampler could be used.
     201.4  When more than one laboratory 1s Involved 1n the
            analysis of the various parameters, the sample
            should, 1f at all possible, not be divided in
            the field but rather at the contractors'
            laboratory.  For purpose of this program the
            composite will be divided Into four parts, one
            part for metals analysis, one for pesticides and
            PCB's, one for SC/MS compounds and one for the
            classic parameters.
     201.5  Blend the composite sample to provide a
            homogeneous mix-Sure Including a representative
            suspension of any solids 1n the container.  No
            specific method 1s required; hand stirring with
            clean glass or Teflon rods, mechanical paddles
            or magnetic mixing with Teflon coated stirring
            bars may be used.  Metal mixing devices may not
            be used.
202  Preservation of Extractable Orqanics
     202.1  Do not add any preservative to the sample; just
            seal and prepare for shipment.  All samples mist
            be carefully identified using labels supplied by
            ESD.  Indicate on the label whether the sample

-------
            1s a raw discharge or treated effluent as
            shown.  If sample 1s to be run on the plasma
            unit, only Indicate so at base of tag.
203  Field Blank Procedure for Automatic Samplers
     203.1  Blank Water - Blank water must be as free from
            organic Interferences as possible.  The
            analytical laboratory should supply this water
            1n bulk glass containers (minimum of five
            liters) for field use.  The supplying laboratory
            shall analyze the blank water to determine the
            organic background that may be present.
     203.2  Procedure - All parts of the sampling system
            must be scrubbed with hot detergent water and
            thoroughly rinsed with tap water and blank water
            prior to use.  Further rinsing with methylene
            chloride is required when parts permit, i.e.,
            are not susceptible to dissolution by the
            solvent.  (Note:  Tygon plastic tubing is a
            source of phthalate ester contamination.  Where
            Its use 1s required, i.e., 1n the peristaltic
            pump, the length must be kept as short as
            possible.)  Teflon is acceptable and may be used
            in other parts of the sampling system as in
            intake lines.  In the field, pump two liters of
            blank water through the sampling line and pump
            tubing and discard.  Then pump three  liters of
                  JUT

-------
            blank water through the system and collect as a
            blank 1n a 1 -gall on sample bottle that has been
            prepared as described below.  Seal the bottle
            with a Teflon lined cap.  Immediately 1ce the
            blank (4°C) and maintain at 4°C during the
            transport and storage prior to analysis.
     203.3  Composite Container - Prepare narrow-mouth glass
            sample bottles for use by washing with hot
            detergent water and thoroughly rinsing with tap
            water and blank water.  Heat the bottles at
            400°C 1n a muffle-furnace or dry heat
            sterilizer for 30 minutes or alternatively;
            rinse with methyl ene chloride and air dry at
            room temperature protected from atmospheric or
            other sources of contamination.  Caps for the
            bottles must be lined with Teflon which has been
            solvent rinsed as above.
204  Volatile Orqanlcs (Purqeables)
     204.1  Containers - Use 45 to 125 ml screw cap glass
            vials with Teflon faced sHlcone septa:
            204.1.1  Vials - Pierce #13074 or equivalent.
            204.1.2  Septa - Pierce #12722 or equivalent.
     204.2  Wash the bottles, septa, and caps with hot water
            and thoroughly rinse with tap water and blank
            water.  Heat the bottles and septa at 105°C
            for one hour, cool to room temperature 1n an
                    J7T-

-------
       enclosed contaminant-free area.  When cool, seal
       bottles with septa (Teflon side down) and screw
       cap.  Maintain the bottles 1n this condition
       until Just prior to filling with blank water or
       sample.
204.3  Collect duplicates 45-125 ml samples each time
       samples are collected.  Two blank water samples,
       sealed 1n 45 ml vials, are to accompany the
       sample bottles during shipment to and from the
       sampling site.  If preservation for residual
       chlorine is to be used, collect four samples
       during each sampling period.  Two should be
       preserved and two not preserved.  Two preserved
       and two non-prssflarved blanks are to be provided-.
204.4  Filling and Sealing Bottles - Slowly fill each
       container to overflowing.  Carefully set the
       container on a level surface.  Place the septum
       (Teflon side down) on the convex sample
       meniscus.  Seal the sample with the screw cap.
       To  ensure that the sample has been properly
       sealed, invert the sample and lightly tap the
       Hd on a solid surface.  The absence of
       entrapped air bubbles indicates a proper seal.
       If  air bubbles are present, open the bottle, add

-------
            additional sample, and reseal.  The sample must
            remain hermetically sealed until 1t Is analyzed.
205  Preservation of Volatile Orqanlcs
     205*1  Preservative (sodium thlosulfate or sodium
            bisulfite) 1s used to stabilize samples
            containing residual chlorine.  The production of*
            chloroform and other haloforms continues 1n such
            samples 1f they are not stabilized.  Waste
            streams that have been treated with chlorine
            should be tested on-s1te to determine whether or
            not preservative 1s needed.  If preservation 1s
            required, collect both preserved and non-<
            preserved samples,  wrap the samples with water
            proof packing material, place In an Insulated
            chest and 1ce at 4°C.  Maintain at 4°C
            during transport and storage prior to analysis.
206  Grab Samples for Volatile Organ1cs (Purgeables)
     206.1  Collect grab samples (minimum of one per day)
            for the analysis of volatile organlcs.  Collect
            samples from the raw process discharge, the
            treated effluent, and the treated effluent after
            chlorlnatlon, when chlorfnation 1s practiced.
            It 1s recommended that the samples be collected
            from mid-channel at mid-depth.  Samples should
            be collected at a turbulent, well mixed section
            of the channel.
                      ML- 37

-------
References
1.  Eichelberger, J.W., L.E. Harris, and W.L.  Budde,  "Reference
    Compound to Calibrate Ion Abundance Measurements  1n Gas
    Chromatography—Mass Spectrometry Systems," Anal. Chem. 47.
    995-1000 (1975).
2.  McNalr, H.M., and E.J. Bonelli, "Basic Gas Chromatography,"
    Consolidated Printing, p. 52, Berkeley, CA, 1969.

-------
                    TABLE 1
   DFTPP Key Ions and Ion Abundance Criteria
Mass               Ion Abundance Criteria

 51                30-60% of mass 198

 68                less than 25! of mass 69
 70                less than 22 of mass 69

127                40-60* of mass 198
    «
197                less than 1% of mass 198
198                base peak, 1002 relative abundance
199                5-9* of mass 198

275                10-30X of mass 198

365                greater than 12 of mass 198

441                less than mass 443
442                greater than 402 of mass 198
443                17-23S of mass 442
                  :&:

-------
                    TABLE 2
       BFB Key  Ions and Abundance Criteria
 Mass               Ion Abundance Criteria

 50                 20-40* of mass 95
 75                 50-70* of mass 95
 95                 base peak, 100* relative abundance
 96                 5-9* of mass 95
173                 less than 1* of mass 95
174                 70-90* of mass 95
175                 5-9* of mass 95
176                 70-90* of mass 95
177                 5-9* of mass 95
             OF- 2a

-------
                 TAIUNG FACTORa ~
                                 A8
Example calculation: Peak HeightsDE = 100 mm
                  10% Peak Height a BO *10 mm
                  Peak Width at 10% Peak Height sACs 23 mm
                        A8 = 11 mm
                        SC =* T2 mm

                  Therefore: Tailing Factors— s1.1
                                        11
          Rgure 1. Tailing factor calculation
            3ZT-

-------
                            Appendix A
                    Storage of Final Extracts

Each extract 1s to be washed  out  of  Its  container into a 10 ml
glass ampul and brought to 5  ml * 1 ml.   Methylene chloride is
the solvent for the base-neutrals and acids, hexane for
pesticides.  The cooled ampuls (-20°C)  are to  be sealed in a
rounded-off, fire polished manner, i.e.,  no thin sharp peaks of
glass that are easily, broken  on handling and shipping.  After
sealing the ampuls, put an indelible mark at the solvent level.
Securely attach a label or tag that  gives:
         Type of fraction (base-neutral,  etc.)
         Industrial category
         Nane (of plant, city and state)
         Specific source of stage of treatonent
         Date sampled
         Date sealed
         Name of contractor and analytical laboratory
         Final volume of extract.
Wrap the ampuls in packing material  to  prevent breakage and mail
or ship them postpaid at ambient  temperature.   When the samples
are safely in ampuls, the remainder  of  the composite sample may
be discarded.

-------
                            Appendix  8
              Archiving SC/MS Data on Magnetic  Tape

    As the analyses are completed, transfer GC/MS raw data to
magnetic tape as described below.  All raw GC/MS data is  to be
retained on 9-track magnetic tape.  When  a tape is filled,
deliver the tape to the Environmental Protection Agency for
storage and later evaluation.  The tape format  is:
         Type - 9-track, 800 BPI, 2400 foot reels
         Record length - 80
         Block size - multiple of 80  (specify)
         Code - ASCII or EBCDIC
The data format must be approved by EPA.  At  the present  time
there are formats used by commercially available equipment that
have been approved by EPA.  Before submitting tapes  to EPA,  a
sample data tape with appropriate documentation should be
submitted for approval of the format  and  content.
    Each GC/MS run on tape should contain in  the following
header information at minimum.
    1.   Identifying run number
    2.   Data and time of sample run
    3.   Run description.  Include GC column  used, temp,  program
         (e.g. 70-270- 8), concentration  factor of extract,  run
         type (e.g. Base/Neutral, Acid, VQA,  Direct  Aqueous
         injectton).
    4.   Scan time in seconds.
                                  32'

-------
    5.   Mass range scanned.
Each scan within each run should contain at minimum:
    1.   Identifying run number
    2.   Scan number
    3.   Date
    4.   Time of scan (time from start of run).
    5.   Total Intensity of scan (not normalized).
    6.   If Intensities are normalized, the mass and raw
         Intensity of the base peak for the scan.  Intensities
         may be normalized or unnormallzed.  Normalized values
         must have.a dynamic range of at least. 1000;
         unnormallzed values must reflect the  actual absolute
         Intensity value.  Masses may be nominal Integer values.
Each magnetic tape reel shall be accompanied by Table of MS
response ratios and a descriptor sheet that Includes the
following information for each run.  See GC/MS Data Transmittal
Form, this Appendix.
    1.   EPA sample number
    2.   Laboratory sample number
    3.   Fraction type  (Base/Neutral, Acid, VGA, Blank, Standard
         Runs, etc.)
    4.   Industrial category
    5.   Specific source or stage of treatment
    6.   Date  sampled
    7.   Sampler  (Contractor or EPA Region)
    8.   Data run

-------
9.   File number on-tape
10.  File name on tape
11.  Comments • to Include any deviation from methods 624 or
     625.

-------
                GC/MS DATA THANSMITTAL FORM
1
2
3
4
5
S
7
3
9
10
11
12
{?•«—*-












u.ruu,












Tv«* ''*c»w












ClMWV












^irz:












OIK












CM«MI« *












On*












                GC/MS DATA TRANSM1TTAL FORM
   (ft
 1
 2
 3
 4
 5
 S
 7
 3
 9
10
11
12

-------
             Quality Assurance and Quality Control Procedures for
             Verification Analysis For Organic Priority Pollutants
110 Scope
    111   These procedures are provided for use by laboratories performing
         verification analyses using EPA Methods 624 and 625.  Verification
         analyses require quantitative determinations with known precision
         and accuracy, so that these data will be acceptable for use 1n
         developing effluent limitation guidelines.  To provide data with
         this degree of reliability, a strong quality assurance and quality
         control program 1s presented.  These procedures are designed to
         produce data with known precision and accuracy so that a
         determination of confidence can be placed in the data.  Quality
         Assurance (QA) 1s the total program for assuring the reliability of
         the monitoring data.  Quality control (QC) 1s the routine
         application of procedures for controlling the measurement process.
    112  Initially, the methodology must be validated for each Industrial
         subcategory.  The requirement for validation of each subcategory is
         based on the unique nature of the wastewater associated with most
         of the subcategories.  Since the effluent from treatment is to be
         used for setting control limits, 1t will be used to develop Initial
         validation data for the method prior to routine sampling and
         analyses.  Based on screening data, a particular subcategory may
         not require verification analyses of all four fractions (Volatile*,
         Base/Neutrals, Adds, and Pesticides) or for all of the individual
         Consent Decree compounds, in which case, the method requires
         validation only for the fraction or the selected compounds of
         interest.

-------
    113   The  results  of  the  validation analyses will  be used to provide
         Information  with which  to  judge  a laboratory's ability to Interpret
         and  i rip lament the method for each Industrial subcategory.  Intlal
         QC limits  for precision and accuracy will  be established using
         these  results,  and  then used in  subsequent analyses as control
         limits.  A numerical  example 1s  given in Appendix C.
    114   After  the  method  1s validated for each subcategory and routine
         analysis begins, continuing QA/QC will be required to ensure that
         the  subsequent  analyses are within the established control limits.
    115   Prior  to developing initial method validation and a continuing
         quality  control program, the analyst (individual or group if team
         approach 1s  used) must  demonstrate the ability to perform the
         required analyses.   If  a laboratory has not established precision
         and  accuracy criteria for  clean  water, the laboratory must perform
         replicate  analyses  of clean water as prescribed in section 121.
         Such data  must  be presented.to the responsible EGO project officer
         prior  to the beginning  of  the verification phase.
120 Routine Quality Assurance and Quality Control
    121   Preliminary Clean Water Precision and Accuracy
         121.1   Before  any work  is  begun on actual field samples, a labora-
                tory must  demonstrate its ability to properly perform the
                11quid-11quid extractions, the gas purge extractions, and
                the required chromatography.  Clean water spikes  are ana-
                lyzed to demonstrate the laboratory's ability to  implement

-------
       Methods 624 and 625, and to establish the baseline precision
       and accuracy criteria for the method 1n that laboratory-
121.2  Procedure:
       121.2.1  Prepare "organic-free* water for use 1n determining
                preliminary precision and accuracy according to the
                procedures given 1n Methods 624 and 62S.
       121.2.2  Spike four replicates of clean water with each
                compound of Interest at a concentration
                approximately equal to 10 times the limit of
                detection.  In addition, spike all purgeable
                allquots with a minimum of three surrogate
                standards at a level of 100 ug/1.  For extractable
                organlcs, each replicate must be one liter; each
                purgeable sample requires at least 100 ml.  Oo not
                dose purgeables with more than 20 ul of an
                alcoholic standard per each 100 ml of water.
                Analyze spiked solutions according to Method 624 or
                625.
121.3  Precision - For each parameter, use the resulting observed
       values of the spikes (0-j, 02, Oj, and 0^) to
       calculate the standard deviation (S) of the replicates
       according to Equation 1.
                           n(n-l)
       Where:
                n * number of replicatas

-------
     121.4  Accuracy - For each parameter, use the resulting observed
            values of the spikes (0-j, 02, 03, and 04) to
            calculate the mean percent recovery (P) of the method
            according to equation 2.
                                n
            Eq. 2      P
                               nr
            Where:
                     n * number of replicates
                     T » true value of the spike
     121.5  The precision and accuracy data shall be documented for the
            record as evidence that the laboratory can properly perform
            the extractions and chroraatography essential for methods 624
            and 625.
122  Method Blank - The method blank is defined as an appropriate volume
     of "organic-free" water which has been processed exactly as the
     sample (including glassware, reagents, solvents, etc.).  Reagents
     or solvents having background levels that interfere with the
     compounds to be determined must be purified and shown to be
     acceptable or replaced with some that are acceptable prior to
     proceeding with analyses.  Problems encountered and corrective
     actions taken shall be documented and reported for the record.
     122.1  For the extractable fractions (Base/Neutral, Acid, and
            Pesticide) the method blank requires extraction of 1-11 tar
            "organic-free11 water.  A method blank must be extracted for
            each set of field samples extracted at a given time (at
            least one method blank per 20 field samples analyzed) and

-------
       whenever a new source of reagent or solvent 1s introduced
       Into the analytical scheme.  The method blank can be
       screened by SC-FIC.  Analysis by GC-MS 1s required only if
       6C-FID analysis of the field blank gives any peaks larger
       than the internal standard peak.
122.2  For the volatile fraction, 5 ml of "organic-free* water
       should be analyzed by the purge and trap methodology only if
       positive interferences are noted during the analysis of a
       field blank.  If positive interference still occurs, repeat
       the method blank analysis.  If interference persists,
       dismantle the system, thoroughly clean all parts that
       contact the sample, purge gas and carrier gas.  Replace or
       repack the sorbent trap and change purge and carrier gas.
123    Field Blank - The field blank is defined as an appropriate
                                                   341*"
       volume of "organic-free" water which has been/\to the
       sampling site and back to the analytical laboratory in a
       container and bottle identical to the type used to collect
       the samples.  Field blanks and samples must be shipped' in
       separate containers.  When received in the lab, the field
       blank is dosed, extracted and concentrated as if it were an
       actual sample.
123.1  For the extractable factors (Base/Neutral, Acid, and
       Pesticide), the field blank may be screened by GC-FTD or
       GC-eC (Pesticides) only).  Full GC-MS analysis is to be
       performed if the screening analysis gives any peaks larger
       than the internal standard peak.

-------
     123.2  For the volatile fraction, analyze a 5-ml field blank after
            each sample analysis.  Follow the guidelines in 121.2 if
            positive interferences are noted during the analysis of a
            field blank.
124  GC-MS Calibration Check
     124.1  For the Base/Neutral-Acid or Pesticide fraction, daily
            analyze decafluorotriphenylphosphine (DFTPP) by injection
            into the GC inlet.  See Reference 1.  (DFTPP must also be
            run after any mass spectrometer tuning is done).  This
            calibration check may be done in conjunction with Section
            125.  The requirement is that 50 nanograms of DFTPP injected
            must meet the specifications prescribed.  Table 1 lists the
            DFTPP key ions and ion abundance criteria.
     124.2  For the volatile (purgeable) fraction, analyze
            4-brcrao-fluorobenzene (BFB) daily by direct injection into
            the GC/MS.  The requirement is that 20 nanograms of BFB
            injected must meet the prescribed specifications.  Table 2
            lists the BFB key ions and ion abundance criteria.  This
            calibration check may be done in conjunction with Section
            125.
125  GC Performance Check
     125.1  SP-2250 column - At the beginning of each day that
            base/neutral or pesticide fraction analyses are to be
            performed,  inject 100 nanograms of benzidine (for
            base/neutral analyses) or aldrin (pesticide analyses) either
            separately  or as part of  a standard mixture that may  also
            contain 50  ng of DFTPP.   Performance for base neutrals

-------
            1s acceptable 1f the tailing factor for benzidine 1s less
            than 3.   Performance for the pesticides 1s acceptable 1f the
            tailing  factor for aldrln 1s less than 2.  Calculation of
            the tailing factor 1s given 1n Reference 2 and described In
            Figure 1.  Retain all plots of performance evaluation as
            evidence of valid performance.
     125.2  SP-1240  DA Column - At the beginning of each day that acid
            fraction analyses are to be performed, Inject 100 nanograms
            of pentachlorophenol either separately or as part of a
            standard mixture that may also contain DFTPP.  Performance
            is acceptable if the tailing factor is less than 5.
            Calculation of the tailing factor is given 1n Reference 2
            and described in Figure 1.  Retain all plots of performance
            evaluation as evidence of valid performance.
     125.3  Carbopak C with Carbowax 1500 or Carbopak B with SP-1000
            column.   At the beginning of each day that purgeables
            analysis are to be performed, inject 20 nanograms of BF3
            either separately or as part of a standard mixture.
            Performance is acceptable if the tailing factor is less than
            2.  Calculation of the tailing factor is given in Reference
            2 and described in Figure 1.  Retain all plots of
            performance evaluation as evidence of valid performance.
126  Performance Evaluation - Ones each 20 working days; dose, extract,
     and analyze a performance evaluation sample (available from
     EMSL-C1ncinnati).  Retain the results as evidence of valid
     performance.

-------
    127  Saving Extracts -If the final extract 1s to be retained, or shipped
         to another location, follow the appropriate guidelines given 1n
         Appendix A.
    128  Archiving GC/MS Data - If the SC/MS data are to be archived on
         magnetic tape, follow the guidelines In Appendix 8.
130 Method Validation
    131  Extractable Orqanlcs (Base/Neutrals-Acids and Pesticides) - The
         following procedures are to be applied, separately, to samples
         being analyzed for the Base/Neutral-Acid group of compounds and to
         samples being analyzed for the Pesticides group.  The analyses
         shall be performed according to the procedures given in Method
         625.  The validation studies must be performed under the same
         conditions ordinarily applied to the samples of a given
         subcategory.  That is, if separatory funnels are routinely used for
         extraction of the samples, the study must be conducted using
         separatory funnels.  If continuous extractors are used for routine
         analysis of the subcategory, the validation study must be performed
         using the same type of continuous extractors.
         131.1  Sample pretreatment - The laboratory will receive a 24-hour
                composite sample of adequate volume to carry out the
                validation study and one field blank taken as described  in
                the sampling protocol on the same day from the same source.
                Mix composite sample and withdraw a 1-liter aliquot for
                analysis.  Vigorously mix the sample by some type of
                stirring device.  Withdraw aliquots, while stirring,  into  a
                1-liter graduated cylinder, using a siphon made of glass or

-------
       Teflon.   Measure and record the volume.  Transfer the
       aliquot  to a 2-11ter separatory funnel or continuous
       extractor for spiking.  Initially analyze a 1-11ter aliquot
       of the sample to determine the sample background so that
       proper spiking levels can be selected for 131.2.1.  The
       remainder of the composite sample should be stored at 4°C
       until  the validation study 1s begun.  At the same time,
       analyze  a 1-11ter aliquot of the field blank.  Choose three
       levels of priority pollutant spikes to cover the expected
       concentration range of the samples 1n the subcategory.
131.2  Preparation of Allquots for Validation -Study - Withdraw
       twelve 1-11ter allquots from the stirred composite sample as
       described 1n 131.1,  Separate Into three groups of four.
       131.2.1   Spiking of AHquots - Spike two allquots of each
                     »
                group with surrogate standards only.  The other two
                allquots are spiked with surrogate standards plus
                the priority pollutant standards of Interest at one
                of the concentration levels (See Figure 2}.  Repeat
                this process for each group of allquots.  Select
                the three spiking concentrations for the priority
                pollutants based on the results of the background
                analysis obtained 1n 131.1.  If the Initial
                background level for a particular priority
                pollutant 1s X, select the three spiking levels to
                give final concentrations of 2X, 10X, and 100X.  If
                X equals 15 ug/1, dose with 15, 135, and 1435 ug

-------
         per liter.   This gives final concentrations of
         priority pollutants of 30, 150, and 1500 ug/1.
         Spike each  1-11ter replicate with each surrogate at
         a level  of  100 ug/1.  (Note:  Consideration should
         be given to the water solubility of the compounds
         being spiked when selecting the spiking
         concentration levels.)
131.2.2  Prepare spiking standards 1n concentrations such
         that no more than 5 ml of spiking solution 1s added
         for each liter of sample.  This will ensure that
        •the solubility of the standard In water will not be
         significantly affected by the added organic
         solvent.  Add the spiking solution to the sample
         allquots 1n the separatory funnel using a transfer
         plpet.  After adding the spikes, thoroughly mix the
         samples and after one hour at room temperature
         proceed with the extraction. ' If continuous
         extractors are used, It may be necessary to spike
         the 1 liter allquots before they are placed Into
         the extractor.  Place the aliquot in a separatory
         funnel or a clean bottle, spike, and transfer to
         the extractor.  The bottle must be rinsed with
         solvent.  Malt an hour and begin the extraction.
         (Note:  Extractions of all  samples must begin
         within 48 hours of receipt  in the laboratory.)

-------
131.3  Use of the Data from Spiked Samples 1n Analyses - The data
       obtained from the determination of priority pollutants are
       used to calculate the precision and accuracy of the method
       and to establish control limits for the Individual compounds
       of Interest.  Surrogate spikes are added to every sample  to
       provide quality control on every sample .by monitoring for
       matrix effects and gross sample processing errors.  The
       surrogate 1s not used as an Internal standard for
       quantification purposes.  Suggested surrogate spikes are
       given 1n Section 6.4 of Method 625.  If verification 1s
       needed for only one fraction, only the surrogates for that
       fraction shall be added.
131.4  Extract and analyse all allquots as directed 1n Method 625
       or other appropriate EPA methods.
131.5  Calculation of Precision and Accuracy - The precision of  the
       method may be calculated from the data obtained during the
       validation study.  There are three spiked concentration
       levels of priority pollutants as outlined 1n Figure 2. The
       method precision for the background level of priority
       pollutants occurring in the sample may be calculated from
       the three pairs of replicate allquots which are spiked only
       with surrogate standards (Al, A2; 81, 82; Cl, and C2.  See
       Figure 2).  The precision and accuracy for the surrogates
       may be calculated from all twelve replicatas sines the
       spiking level is constant for both sets of six samples.   The
       precision and accuracy for the surrogates may be calculated

-------
for each set of six samples  if  there  1s  an  effect due to the
added priority pollutant spike.   Similarly,  the  precision
and accuracy at each spike level  of the  priority pollutants
may be determined from the two  replicate aliquots that
received that spike (01, D2;  El,  E2;  Fl. and F2  See Figure
2).
a.  Precision
Calculate the range (R) for  each  pair of replicate aliquots
1e.t duplicate analyses, according to equation 3:
      Eq. 3   R
Where:
     X-i and X- are each an analytical result from two
     replicate aliquots.
The concentration level related to R  can be represented as
in equation 4.

       Eq. 4    I -  (Xt + X2)
                         2
Where:
        X is the mean of the duplicate analyses, X-j and
        v
For any group of n duplicate analyses that  are considered
similar to each other, their ranges  (R^) and means (X.),
where 1 * 1 to n, can be used to  estimate the critical
difference (Rc) between similar future duplicate analyses
or any specific concentration leva!  (C). Calculate R  as
shown in aquation 5.
                izr       r

-------
    Eq.  5
                                 1-1
From these data develop a table of RC values for various
concentration (C) values that span the concentration range of
interest.  Use these Initial critical difference values to
judge the acceptability of succeeding duplicate results
generated under the same conditions.  Revise and update as
additional duplicate data becomes available.  When more than
15 pairs of duplicates are available within any specific
concentration level C, RC should be calculated directly
from the average range of these duplicates alone.  Equation 5
reduces to
                                  n
                                 2Z  RI
   b.   Accuracy for Priority Pollutants
        Calculate the background concentration of
        priority pollutants occurring 1n each of the
        field composite samples.  The calculation is
        similar to equation 4, but there are six
        pieces of data (AT, A2, Bl, 82, Cl, C2)
        available for this calculation as shown in
        Figure 2.  Therefore, the calculation is as
        shown in equation 6.
                             6
          Eq. 6     I  »  ( £ XJ
                             1-1
                                ,
                                '

-------
 Where:
       X* 1s the meanjXj, 1=1 to 6 are the analytical
       results for the six 1-Hter allquots of a single
       composite sample spiked with surrogates only.
       Calculate the recovery of each priority pollutant
       1ir each of the 1-11 ter allquots spiked with
       priority pollutants (01, D2. El, E2, FT and F2)
       according to equation 7:

           Eq. 7    P - 100 (2 - 7)
Where:
       P 1s the percent recovery of the spike
       2 1s the analytically determined
       concentration of the priority pollutant  in
       the spiked aliquot
       X is the mean background concentration of
       the priority pollutant and
       T is the true value of the spike.

Determine the percent recoveries for each priority
pollutant at all of its concentration  levels.   If
there is no significant difference between the
percent recoveries for the various concentration

-------
   levels, all n of the percent recovery values may be
   treated together as 1n equations 8 and 9.   If  some
   of the percent recovery values are significantly
   different, each group of similar percent recoveries
   must be treated Independently to develop Its own
   characteristic mean percent recovery (P) and Its
   associated standard devlalon (S }.
                n
    Eq. 3  7 «
Where:
    P 1s the mean percent recovery
                              •
    P 1s an Individual percent recovery value
    n 1s the number of observations  at this
      concentrator! level

-------
           ».»
                               n(n-l)
           c.  Accuracy for Surrogates
           Proceed exactly as with priority pollutants  in  131.5b  above,
           keeping the following two differences  in mind:  there  is  no
           background concentration and there  are six sets of duplicate
           analyses for the surrogate spikes;  three sets spiked with
           priority pollutant  (D1, 02; £1, E2;  FT, F2)  and three  sets
           without (Al, A2; Bl, 82; C1, C2), see  Figure 2.  Calculate
           the percent recovery as shown  in equation  10.

                     Eq.  10  P - 100 Z

                 Where:
                      P is the percent recovery of the  surrogate  spike.
                      Z is the measured value  of  the  surrogate  spikes
                      in  the aliquot.
                      T is the calculated or true value of the
                      surrogate spikes added to the sample.
           Calculate the  mean  percent recovery (P) and  the standard
           deviation  (S ) of the percent  recovery of  the surrogate
           spikes in  all  of the sample aliquots according  to  equations
           3 and 9.
132  Volatile Organics  (Purgeables)
     132.1 All  individual grab  samples collected  during the  specified
            time period  shall be  composited in  the laboratory,  spiked
            and analyzed  as  soon  as possible.   -Significant  losses  of

-------
       some of the purgeable priority pollutants will occur  in  1  to
       2 days.  However, in a nonreactlve matrix, most are stable
       up to 14 days.  The method validation study requires  twelve
       5-ffll allquots from each composited sample.  Carefully
       transfer the remaining volume of composited sample back  to a
       clean vial or vials and seal with no headspace as done when
       collecting a sample.  These should be held at 4°C until  it
       1s determined that there 1s no further need for the sample.
       Figure 3 summarizes the validation study for volatile
       organlcs.
       Caution:  Prepare only as many sample aliquots as can be
       analyzed in the working day. .This may mean that each of the
       three concentration levels will be analyzed on different
       days.
132.2  Pretreatnent of Grab Samples - Individual grab samples
       should be composited according to the following procedure:
       a.   Composite only grab sanples of equal volume.
       b.   Carefully pour the contents of all individual grab
            sanples collected from a given source during the
            specified time period into a 1000-flil round-bottom flask
            which Is chilled 1n a wet ice bath.
       c.   Stir the mixture gently with a glass rod for
            approximately one minute while in the ice bath.
       d.   Carefully fill 13 clean 40-ml vials or three 120-*nl
            vials and four 40-ml vials with ccmpositad sample.
       e.   Take one 40-
-------
       f.   Store the remaining vials at 4°C until the validation
            study 1s begun.
132.3  Spiking levels for priority pollutants and surrogate
       standards - The spiking levels of the priority pollutants
       are determined by the background (X) 1n the samples.  The
       low level spike will give a final concentration that 1s 2
       times the background level.  The Intermediate and high level
       spikes will give final concentrations that are 10 and 100
       times the background level.  Concentrations 1n excess of
       1000 ug/1 are likely to flood the gas chromatographic
       column.  Therefore, the total concentration (background plus
       spike) of each Individual priority pollutant should not
       exceed 900 ug/1.  Even at this level the solubility of the
       compounds in the sample must be considered.  The spiking
       level for all surrogate standards should be 100 ug/1.
132.4  Spiking Procedures
       132.4.1  Preparation of Spiking Standards - Prepare
                methanolic stock standard solutions of the priority
                pollutants and the surrogate standards according to
                the directions given in Method 624.
                From the methanolic primary dilutions prepare
                secondary aqueous spiking mixtures of the surrogate
                standards so that 20.0 ul of the primary standard
                solution, diluted to 50.0 ml 1n organic free water
                will permit adding 5 ul of the resulting solution
                to the 5 ml sample giving the desired surrogate

-------
         concentration level of 100 ug/1.
         Prepare spiking mixtures of the priority pollutant
         standards 1n methanol so that 20.0 ul of the
         solution added to 100.0 ml of sample will give the
         desired concentration levels.
132.4.2  To minimize the solubility effect of methanol on
         the constituents to be measured, do not Inject more
         than 20 ul of spiking solution per 100 ml of
         sample.  Never use a pipet to transfer samples or
         aqueous standards that are to be analyzed for
         volatile purgeable compounds.  Transfer samples by
         pouring Into the receiving vessel.
                                                     *
132.4.3  Spiking the Sample AHquots - Take one of the
         120-ml or 3 of the 40 ml sample allquots from cold
         storage, equilibrate to room temperature, and fill
         a 100 ml volumetric flask to mark with the sample.
         Rapidly Inject 20 ul of the methanoUc solution of
         priority pollutant spiking standard (concentration
         2X) Into the expanded area of the flask below the
         neck.  Stopper and nrix by gently Inverting the
         flask three times.  Fill two 5-ml syringes with
         spiked sample from the flask as directed 1n the
         analytical protocol.  Open the valve of the syringe
         and Inject 5 ul of the surrogate standard spiking
         solution.  Inject the sample aliquot Into the
         purging device and analyze according to Method 624.

-------
                         Take one  of the 40-ral  sample aliquots from cold
                         storage,  equilibrate to  room temperature  and fill
                         two  5-ffll  syringes  with the  sample as directed 1n
                         Method  624.   Spike 5 ul  of  the surrogate  standard
                         water solution  (concentration 100 ug/1)  Into the
                         syringe through the valve giving a final
                         concentration of 100 ug/1.   Inject the sample
                        •aliquot Into the purge device and analyze according
                         to Method 624.   See Figure  3.  Repeat this
                         procedure twice, giving  three sets of analyses of
                         two  samples spiked with  surrogate standards only
                         and  two samples spiked with surrogate standards and
                         priority  pollutants.
         132*5  Calculation of Precision and Accuracy - The precision and
                accuracy for  the purgeable  priority  pollutants and the
                surrogate standards are  calculated as directed for the
                semlvolatHe  solvent extractable  compounds In paragraphs
                131.5a,  b, and c.
140 Continuing Quality Assurance and quality Control
    141  Extractable Organics  (First Day Samples) - The following
         procedures should be applied to the first day samples of the
         pesticides group Base/Neutral -Add group.  They should be carried
         out on each first day composite sample for  each new subcatagory.
         This Includes samples of  the Influent  to treatment, the effluent
         from treatment and the supply water.   An outline diagram for first

-------
day ongoing quailtyassurance samples 1s given in Figure 4.
141.1  Withdraw three 1-1 Her allquots of the composite sample
       according to the procedure 1n 131.1.
141.2  Spiking the Sample AHquots - Spike one of the allquots with
       priority pollutant standards plus the surrogate standards
       and two of the allquots with surrogate standards only.
141.3  Add a spike sufficient to approximately double the
       background concentration of the priority pollutants as
       determined In 131.5b-.  If the original concentration Is
       higher than the midpoint of the calibration curve, then the
       concentration- of the spike should be approximately one-half
       the original concentration.  Surrogate spikes as specified
       1n 131.3 should be added to all three allquots from each
       sample at a concentration level of 100 ug/1.
141.4  Analyze according to Method 625.
141.5  Calculations of Precision and Accuracy
       a.   For the first day samples, calculate the precision of
            the duplicate analyses (X^ and X^) from the two
            l-11ter allquots for the priority pollutants background
            and the surrogate standards.  Calculate the range (R)
            of the results according to equation 11.
                  Eq. 11   R «|X., - X2|
            The concentration of each compound Is represented by
            the mean of the duplicate values.  Calculate the mean
            (X) according to equation 12.
                   Eq. 12   7 » (X7 * X?)

-------
   Refer to the table of critical range values developed in
   131.5a, to find the concentration (C) nearest to X.
   Use this RC to evaluate the acceptability of R from
   Eq. 11.  If R is greater than RC, the system precision
   is out of control and the source of this unusual
   variability should be identified and resoJved before
   continuing with routine analyses.  After correcting the
   source of this unusual variability, reanalyze the sample
   if possible.  Record the results of all duplicate
   analyses and periodically (after 5-10 additional
   duplicate results are obtained), revise, update, and
   improve the table of critical range values.
          »
b. Accuracy for Surrogate Spikes
   Calculate the recovery of the surrogate spikes in the
   duplicates according to equation 13.

            Eq. 13    P • 100 I

   Where:
     P is the percent recovery.
     Z is the analytically determined concentration of the
     surrogate standard spikes.
     T is the true value of the  surrogate standard  spikes
     added in 132.4.3.
   If the percent recoveries are not within the  interval ?
   + 3S  as determined  in 131.5c, the system  should be
   checked for problems.  If problems exist,  they must be

-------
    resolved before continuing with routine analyses.
    Record the recovery of all surrogate spikes and
    periodically (every 5 to 10 additional data points),
    revise, update and Improve the recovery criteria.
c.  Accuracy for Priority Pollutant Spikes
    Using the results obtained from the 1-Htar aliquot of
    composite sample spiked with surrogate standards and
    priority pollutant standards, calculate the recoveries
    of the priority pollutants according to equation 14.

                Eq. 14   P - 100 (Z-X)
    Where:
          P 1s the percent recovery
          Z is the analytically determined
          concentration, of the priority pollutant
          spikes
          T is 'the true value of the priority
          pollutant spikes added in 132.4.3, and
          X is the mean concentration of the priority
          pollutant background determined by equation
          9.
    If the percent recovery is not within the interval
    of P *, 3S , as determined in 131.5b the system
    should be checked far problems.  If problems
    exist, they must be resolved before continuing
               HUT-

-------
                with routine analyses.  Record the recovery of all
                spikes and periodically revise, update, and
                Improve the accuracy criteria.
142  Extractable Organics - (Subsequent Day Samples) - The
     following procedures should be applied to each subsequent day
     sample of the Base/Neutrals-Acids group and the pesticide
     group.  A flow diagram for each subsequent day ongoing
     quality assurance samples 1s given in Figure 5.
     142.1  Withdraw a. one-liter aliquot as directed in 131.1
     142.2  Spike the aliquot with surrogate standards at a
            concentration of 100 ug/1.
     142.3  Analyze according to Method 625.
     142.4  Determine the percent recovery of the surrogate
            standards using Equation 10.  If the percent recovery
            is outside the interval 7 + 3S  as determined in
            131.5c, the analytical system should be checked for
            problems.  If problems exist, they must be resolved
            before continuing further sample analyses.
     142.5  A field blank for the day-must be analyzed according
            to Method 625.  If priority pollutants are found and
            quantified, the values for the field blank should be
            noted and reported along with sample results.  If
            significant interference problems occur, the method
            blank must be analyzed to determine if interference
            was introduced in the

-------
               field or the laboratory.  Appropriate action must
               be taken to eliminate the problem before continuing
               with the analysis of routine samples.
143  Volatile Organ1cs (First Day Samples) - Continuing quality
     assurance for the purgeable volatile organics should be
     applied to samples from Influent to treatment, effluent from
     treatment, and supply water.  The following procedures should
     be carried out on each first day composite samples from the
     three sample sources from each subcategory.  An outline is
     given in Figure 4.
     143.1  Composite all relevant grab samples as described in
            Section 132.2.  Prepare six 5-ml allquots for analysis.
     143.2  Spike two allquots with the priority pollutant
            standards at a level twice that determined in Section
            132.5 and the surrogate standards using the procedures
            in Section 132.3 and 132.4.  Spike four 5-flil aliquots
            with surrogate standards only as 1n 132.3 and 132.4.
     143.3  Analyze one of the duplicates spiked with priority
            pollutants and surrogate standards and two of the four
            replicates spiked with surrogate standards only.  The
            remaining spiked aliquots are analyzed only 1f a
            problem is encountered with the analysis of the first
            set of aliquots.

-------
     143.4  Analyze the spiked allquots according to Method 624.
     143.5  Calculate the precision and accuracy as directed for
            the semivolatne solvent extractables as directed in
            141.5.
144  Volatile Orqanics (Subsequent Day  Samples) - The following
     procedures should be applied to each subsequent day sample of
     the volatile organics group.  An outline is given in Figure 5.
     144.1  Composite the relevant grab samples as described in
            Section 132.2.  Prepare two 5-ml aliquots for analysis.
     144.2  Spike both aliquots with surrogate standards only to
            give a concentration of 100 ug/1.
     144.3  Analyze one of the aliquots according to Method 624.
            The other aliquot is analyzed only if a problem is
            encountered..
     144.4  Determine the percent recovery of the surrogate
            standards using Equation 10.  If the percent recovery
            is outside the interval P £ 3 S  as determined in
            131.5c, then the analytical system should be checked
            for problems.  If problems exist, they must be
            resolved before continuing further sample analysis.
     144.5  Analyze a field blank representing the same day that
            the samples were collected.  Follow the guidelines
            given in 142.5.
                             '- 3/2.

-------
200 Sampling Procedures for Verification Phase
    201  These procedures are given as an example for the collection
         of sample*.
    202  Extractafale Organlcs (Base/Neutrals-Adds and Pesticides)
         202.1  Initial QA/QC - Three 2Vgallon samples must be taken
                concurrently over the Initial period of sampling (24
                hrs.) effluent from treatment If both the
                Base/Neutrals-Acid group and the Pesticide group are
                to be determined.  If only one of the two groups 1s to
                be determined, collect two 2%-gallon samples or one
                5-gallon sample.  The samples can be collected using
                two or more'automatic Isco samplers that have the
                sample lines linked together to ensure close replicate
                samples.
                   a.   Cluster the Inlet tubes to ensure that samples
                        are taken geographically close.
                   b.   Link the compositors in a master/slave
                        arrangement, wherein one "master" sampler
                        supplies the sample Interval timing for the
                        "slave" sampler(s).  The Mastar sampler
                        operates 1n a TIME mode and the interval
                        between samples is set on the SAMPLE RATE
                        switch (Model 1580) or the SAMPLE INTERVAL
                        switch (Model 1680).  The "slave" sampler(s)
                        operate in a FLOW mode, with the FLOW
                        MULTIPLIER switch (Model 1580)  set on 1, or

-------
            the SAMPLE INTERVAL switch (Model 1689) set to 1 FLOW
            PULSE.  When the master sampler times out and collects
            a sample, a flow pulse 1s sent to the slave
            sampler(s), causing them to collect samples
            simultaneously with the master sampler.  (Note that
            depressing the MANUAL ADVANCE pushbutton on the master
            sampler will cause the slave sampler(s) to collect a
            sample; therefore; the slave sampler(s) should be left
            1n STANDBY until the master sampler set-up has been
            completed).  No modification to the equipment 1s
            necessary.  Use of two "slave" samplers require a "Y"
            connect cable assembly for the master samples and two
          • sampler-flow meter cables to connect to the slave
            samplers^  .  Samplers should be checked before use
            1n the field to ensure that each collects the required
            volume of sample + 5X,
     202.2  Continuing QA/QC - One 2Jj-gallon sample Is required
            for the continuing QA/QC work.
Available from Isco, Environmental Division, Building
1020, Lincoln Air Park West, Lincoln, Nebraska, 68524.

-------
     202.3  Blanks - One 2-gallon aliquot 1s needed for the preliminary
            QA/QC.  This should be obtained by flushing the organic-free
            water through all Isco samplers used.  One 1-gallon blank  1s
            needed for continuing QA/QC work.
203  Volatile Orqanlcs (Purgeables)
     203.1  Sampling Precautions
            a. Collect 2 replicates from each sample source (I.e.,
               duplicate samples).
            b. Fill the sample bottles 1n such a manner that no air
               bubbles pass through the sample as the bottle 1s being
               filled.
            c. Seal the bottles so that no air bubbles are entrapped 1n
               them.
            d. Maintain the hermetic seal on the sample bottle and keep
               at 4°C until time of analysis.
     203.2  Initial QA/QC - A minimum of 1000 ml (twenty-five 40-flil
            vials) must be collected over the relevant time period at
            each sample point.
     203.3  Continuing QA/QC - A minimum of 200 ml (five 40-ml vials)
            must be collected over the relevant time period at each
            sample point.
     203.4  Blanks - Three 40-flil "trip" blanks (field blanks) are
            required for both the Initial and continuing QA/QC work.

-------
                                  REFERENCES


1.  Elchelberger,  J.W.,  L.E.  Harris,  and W.L.  Budde,  "Reference Compound to
    Calibrate Ion  Abundance Measurements 1n Gas Chromatography-Mass
    Spectrometry Systems," Anal.  Chem.  47.  995-1000 (1975).

2.  McNalr, H.M.,  and E.J. BonelH,  "Basic  Gas Chromatography," p.52,
    Consolidated Printing, Berkeley,  CA  1969.

-------
                  TABLE 1
 OFTPP Key Ions and Ion Abundance Criteria

Mass                     Ion Abundance Criteria
51                       30-60* of mass  198
68                       less than 25 of mass  69
70                       less than 22 of mass  69
127                      40-60* of mass  198
197                      less than IX of mass  198
198                      base peak,  1002 relative  abundance
199                      5-92 of mass 198
275                      10-302 of mass  198
365                      greater than IX of mass 198
441                      less than mass  443
442*                     greater than 402 of mass  198
443                      17-23* of mass  442

-------
                   TABLE 2

   BFB Key Ions and Ion Abundance Criteria



Mass               Ion Abundance Criteria


 50                20-40*-of mass 95

 75                50-70* of mass 95

 95                base peak, 100* relative abundance
 96                5-9* of mass 95

173                less than 1* of mass 95
174                70-90* of mass 95
175                5-9* of mass 95
176                70-90* of mass 95
177                5-9* of mass 95

-------
                 TAIUNG FACTORS —
                                 AS
Example calculation: Peak Height a DE a 100 mm
                  10% Peak Height a BO a 10 mm
                  Peak Width at 10% Peak Height = AC s 23 mm
                        A3 = 11 mm
                        BC»12 mm

                  Therefore: Tailing  Factors—  =1.1
                                         11
          Rgure 1. Tailing factor calculation

-------
                                   Figure 2
                        Summary of Initial Validation
                       Analyses for Extractable Samples
1.  Composite the Replicate Samples.

2.  Withdraw a 1-Hter aliquot.  Store composite  at 40C.
    Separate Into three groups of 4 aliquots each.

3»  Determine X the background concentration of each
    Priority Pollutant.

4.  Withdraw twelve 1-liter allquots from the composite.
5.  a) Dose 2 of allquots with surrogate standards only  at
       100 ug/1.  Label as Al and A2.                          2
    b) Dose 2 allquots with surrogate standard at 100 ug/1
       and priority pollutants to give a concentration
       level of 2X.  Label as 01 and 02.                       2
    c) Analyze Al, A2, 01-, and 02 using Method 625.

6.  Repeat 5a, b, and c.  Label surrogate standards only
    as 81 and 82.  Use 10X level for priority pollutants.
    Label as El and E2.                                        4

7.  Repeat 5a, b, and c.  Label surrogate standards only
    as Cl and C2.  Use 100X level for priority pollutants.
    Label as Fl and F2.                                        4
                                                           Liters     GS/MS
                                                            Used       Runs
                                                       TOTAL  13        13
                             'ML

-------
                                   Figure 3
                        Summary of Initial Validation
                        Analyses for Purgeable  Samples
                                                              5-ffll
                                                            Syringes   SC/MS
                                                              Used      Runs
1.  Composite the grab sample.
2.  F111 13 clean 40-fltI vials or 3 clean  120-ml  vials +
    4 clean 40-nl vials with conposlte, cap,  and store at 4QC.

3.  a) F111 a 5-ffll syringe from one 40-nil vial.                 1
    b) Analyze and determine X the background concentration             1
       of each priority pollutant.

4.  a) F111 two 5-ffll syringes from one 40-flil  vial               2
    b) Dose with surrogate standards at 1QO ug/1.
       Label as Al and A2.
    c) Analyze Al and A2 using Method 624.                               2
    d) Fill a 100 ml volumetric to mark using one  120-ffll
       or three 40nnl vials.
    e) Dose with 20 ul of priority pollutants to give a
       concentration of 2X.
    f) Fill two Snnl syringes from the 100-ml volumetric.      2 .
    g) Dose each syringe with surrogate standards  at  a
       concentration of 100 ug/1.  Label  as 01 and 02.
    h) Analyze 01 and 02 using Method 624.                                2

5.  Repeat 4.  Label surrogates only as B1 and 82.  Use 10X
    level for priority pollutant.  Label  priority  pollutants
    plus surrogate standards as £1 and E2.                      4         4

6.  Repeat 4.  Label surrogates only as C1 and C2.  Use 100X
    level for priority pollutants.  Label priority pollutants
    as F! and F2.                                               4        	4

                                                      TOTAL    13         13

-------
                                   Figure 4
                     Summary of Ongoing Quality Assurance
                            for First Day Samples

                                                              Liters   GC/MS
Extractables                                                   Used     Runs

1.  Composite the Sample.

2.  a) Withdraw three 1-liter aliquots.                          3
    b) Dose two aliquots with surrogate standards
       only at 100 ug/1.
    c) Dose one aliquotvith surrogate standards and
       the priority pollutants of interest to give a
       concentration of 2X, Section 131.5.
    d) Analyze using Method 625.                                           3
                                                 TOTAL          "I       "T
                                                                5 ml
                                                              syringes  GC/MS
Purgeables                                                      Used      Runs

1.  Composite the Sample.

2.  a) Fill 4 clean 4-ml vials or 1 clean  120-ml vial
       •*• 1 clean 40nnl vial with composite.  Store at 4°C.

3.  a) Fill four 5-ml syringes from one 40-ml vial.              4
    b) Dose each with surrogate standards  at 100 ug/1.
    c) Fill a 100 ml volumetric to mark using the 120-ml vial
       or three 40-flil vials.
    d) Dose with 20 ul of priority pollutants to give a
       concentration of 2X, Section 132.5.
    e) Fill two 5-ml syringes from the 100 ml volumetric.        2
    f) Analyze two 5-ml syringes containing surrogate
       standards only and one 5-ml syringes containing surrogate           3
       standards and priority pollutants using method 624.

                                                         TOTAL  ~T~      "T"

-------
                                   Figure 5
                          Suanary of Ongoing Quality
                     Assurance for Subsequent Day Samples

                                                             Liters    GC/MS
Extractables                                                  Used      Runs
1.  Composite the sample.

2.  a) Withdraw a 1-liter aliquot.
    b) Oose with surrogate stand
    c) Analyze using Method 525.
b) Oose with surrogate standards only at TOO ug/1.
                                             Subtotal           r     ^H
                                             TOTAL  (x29 days)  29       29*
                                                               5-ml
                                                             Syringes  6C/MS
Purqeables                                                     Used      Runs

1.  Composite the sample.

2.  a) F111 two 5-fl»l syringes with composite.                   2
    b) Dose with surrogate standards at 100 ug/1.
    c) Analyze one 5-ral sample.                                	     1 •
                                             Subtotal           2        ]
                                             TOTAL  (x29 days)  SB       29*
*Assunring that field blank shows no priority pollutants.
 If field blank were to be analyzed by GC/MS each  subsequent
 day, the total would be 53.
                                      ^3

-------
                                 Appendix A

                          Storage of Final Extracts

Each extract 1s to be washed out of Its container  Into a  10-ml  glass  ampul
and brought to 5 ml £ 1 ml.  Methylene chloride 1s the solvent  for the
base-neutrals and acids,, hexane for pesticides.  The cooled  ampuls  (-20°C)
are to be sealed 1n a rounded-off, fire polished manner,  I.e.,  no thin  sharp
peaks of glass that are easily broken on handling  and shipping. After
sealing the ampuls, put an Indelible mark at the solvent  level. Securely
attach a label or tag that gives:

    Type of fraction (base-neutral, etc.)
    Industrial category
    Name (of plant, city and state)
    Specific source or stage of treatment
    Date sampled
    Date sealed
    Name of contractor and analytical laboratory

Wrap the ampuls 1n .packing material to prevent breakage  and  mall or  ship
than postpaid at snblent temperature.  When the samples  are  safely  1n
ampuls, the remainder of the composite sample may  be discarded.
                                 IT-

-------
                                  Appendix B
                    Archiving SC/MS data on Magnetic Tape
When the analysts are completed, transfer GC/MS raw data to magnetic  tape  as
described below.  All raw SC/MS data 1s to be retained on 9-trade magnetic
tape.  When a tape 1s filled, deliver the tape to the Environmental
Protection Agency for storage and later evaluation.  The tape format  Is:
    Type • 9-tradc, 800 BPI, 2400 foot reels
    Record Length - 80
    Block size - multiple of 80 (specify)
    Code - ASCII or EBCDIC
The data format must be approved by EPA.  At the present time there are
formats used by commercially available equipment that have been approved by
EPA.  Before submitting tapes to EPA, a sample data tape with appropriate
documentation should be submitted for approval of the format and content.

Each SC/MS run on tape .should contain the following header Information at
minimum.
    1.  Identifying run number
    2.  Data and time of sample run
    3.  Run description.  Include GC column used, temp, program
        (e.g. 70-270 £8), concentration factor of extract, run type
      .  (e.g. Base/Neutral, Add, VOA, Direct Aqueous Injection).
    4.  Scan time in seconds.
    5.  Mass range scanned.

-------
Each scan within each run should contain at minimum:
    1.  Identifying run number.
    2.  Scan number
    3.  Date
    4.  Time of scan (time from start of run).
    5.  Total Intensity of scan (not normalized).
    6.   If Intensities'are normalized, the mass and raw  Intensity  of
         the base peak for the scan.
    7.   Mass/Intensity pairs for the scan.  Intensities  may be
         normalized or urmormallzed*  Normalized values must have a
         dynamic range of at least 1000; unnormallzed values must
         reflect the actual absolute Intensity value.  Masses may be
         nominal irrterfer values.
                      t
Each magnetic tape reel shall be accompanied by Table of  MS response ratios
and a descriptor sheet that Includes the following Information for  each
run.  See Figure GC-MS Transnrfttal Form, this Appendix.
    1.  EPA sample number
    2. 'Laboratory sample number
    3.  Fraction type (base/Neutral, Add, VOA, Blank, Standard  Runs,  etc.)
    4.  Industrial Category
    5.  Specific source or stage of treatment
    6.  Date sanpled
    7.  Sanpler (Contractor or EPA Region)
    3.  Data run
    9.  File number on tape
    10. File name on tape
    11. Comments - to Include  any deviation from method  624 or 625.

-------
                GC/US DATA TRANSMITTAL FORM
1
2
3
4
5
6
7
3
9
10
11
12
^P» ^h^^BI












U.«_.












fotfcMMI












l««M>lil












IwalKtaM»«
ln»#t 	 in












OMt
u 	









I
I

Imitm-
CvHXCMi W
t»» liMiimi












DM
•OT












                GC/US DATA TRANSMITTAL FORM
                            MOT!
 1
 2
 3
 4
 5
 6
 7
 3
 9
10
11
12

-------
                                   APPENDIX C

                Numerical  Example of Verification Phase Results


The following 1s an example of the calculations and results of a
.verification study.

         Surrogate Standard
             (ug/1)
                                  Range      Mean         2 Recovery
 Satnole     Added      Found      (Eft. 3)    (Eg. 4)         (Eg. 10)

  Al         95         93                                    98
  A2         95         97          4          95            102
  81         95         96                                   101
  82         95         98          2          97            103
  Cl         95         90                                    95
  C2         95         94          4          92             99
  01         95         99                                   104
  02         95         95          4          97            100
  El         95         89                                    94
  EZ         95         91          2          90             96
  Fl         95         94                                    99
  FZ         95         96          2          95            101

     Critical Difference (Eq. 5)  Rc - 9.8 at 95 ug/1.
     Mean % Recovery (Eq.  8)      ?  •* 99*
     Standard Oev. of P (Eq. 9)   Sp « 3.1        3Sp - 9.4
          Acceptable Range of Recovery  90 to 109*

 During the same verification study, the following data were
 obtained for one of the priority pollutants studied.

                Priority Pollutant                 Percent    Mean
         Back-       '             Range    Mean    Recovery Recovery' Std.Dev.
 Sample  ground Added  Found     (Eg. 3)  (Eo. 4)  (Eo. 7)   (Eg. 8)    (Eo.  9)
                                    2       10  -

                                    3       12.5

                                    1       13.5
                                                      42
                                    1       17.5      50        46         5.9
                                                      82
                                    5      106        86        84         3.3
                                                      96
                                   20     1150        94        95         1.2
X
Al
A2
81
82
Cl
C2
01
D2
El
E2
Fl
F2
X
X
X
X
X
X
X
12
12
12
12
12
12
0
0
0
0
0
0
0
12
12
no
no
1200
1200
12
9
n
n
14
13
14
17
18
102
107
1160
114Q

-------
    Critical Difference    (Eq. 5)  Rc -  6.7 at 12 ug/1
    for priority pollutants         Rc »  3.3 at 17.5 ug/1
                                    Rc » 16.4 at 106 ug/1
                                    Rc • 65 at 1150 ug/1

    Mean Value of X  (Eq. 6)          X « 12.0 ug/1 (background  concentration)

Is there a significant difference 1n the recoveries between  the 100X  and 10X
levels?  Apply a two tailed student's t-test with a confidence  level  of 95*.
v
                                        (3.3)2  . 3.51
                                               2 degrees of freedom
         d • XIOQ - XJQ  -  95-84 « 11

               •  d/Sd • 11/3.51 « 3.70
Since 3.7 1s less than 4.3 (t-value, 0.95, 2 degrees of freedom)  there 1s no
significant difference between the 100X and 10X  levels.  Apply  equations 8
and 9 to the four recoveries for these two levels.  The mean  recovery (eq.8)
1s 89S with a standard devlalon (eq. 9) of 6.5 (3 degrees  of  freedom).   Test
the 2X level against this mean recovery and standard deviation.

         S
-------
For the Surrogate Standards, on day one, the  range between  a and b (Eq.  11) 1s
3, and the mean concentration  (Eq. 12)  1s 91.5  ug/1.   Since the critical
difference 1s 9.8 at 95 ug/1,  this range 1s acceptable.   The recoveries  of the
Surrogate Standards 1n a, b, and c (Eq. 3) are  all  1n  the acceptable range of
^90 to 109S.  Therefore, the accuracy  1s acceptable.

For the Priority Pollutant, on day one, the range  between a and b 1s
acceptable,  (less than 6.7) and the recovery  1s acceptable  (acceptable range
of 28 to 64*).

The following 1s an example of the results for  day 2 to  30  of the 30-day study:
              Priority
            Pollutant (ug/1)
     Day      Found

      2          14'
      3          15
      4          20
      5          11
      6          17
      7          18
      8          10
       9          14
      10          13
      11          12

      12          20

      30          15
Surrogate Standard (ug/1)
Added   Found   % Recovery
 100
 100
  00
 100
 100
 100
 100
 100
 100
 100

 100

 100
 90
 99
100
107
100
100
 75
 92
 93
 94

 95

 95
 90
 99
100
107
100
100
 75
 92
 93
 94

 95
-Update Recovery for SS*

-Not Acceptable
   01sregard
   Results
-Update Recovery for
   SS**
 95 -Give final statistic on
       Recovery of SS.
 *  Includes 20 results,  12 validation,  3 first day and days 2 through 6
      Mean - 99  Std.  Dev.  » 4.6   Acceptable range 85-1132

 ** Includes 25 results;  day 8 result not included.
   •   Mean » 98  Std.  Dev.  » 4.6  Acceptable range 84 -112S

-------
                                   Appendix D
                         Possible  Sources of Standards
 _Aldr1ch Chemical Co., Milwaukee, Wise.;
 -Ana4abs,  Inc., North Haven, Conn.;
 j. T. Baker Chemical Co., PhUHpsburgh, N.J.;
_Chenrfcal Procurement Laboratories, College Point, N.Y.;
•-Columbia Organlcs Catalog A-7, Columbia, S.C.
 Eastman Kodak Co., Rochester, N.Y.;
JCac.Rare & Fine Chemicals, Plainvlew, N.Y.;
• Nanogens International, P.O. Box 487, Freedom, CA  95019
 Pfaltz & Bauer'Chemical Co., Stanford, Conn.;
 RFR Corp., Hope, R.I.; "Chemical Standards for Air-Water-Industry-Foods"
 (1975).
 "Analytical Reference Standards and Supplemental Data for Pesticides  and
 -Other Selected Organic Compounds," EPA-660/9-76-012  (May 1975), Health
 Effects Research Laboratory, Environmental Toxicology Division, Research
 "Triangle Park, 1C.  /I nmfTi ffirnlan VinV rnn ?-u-nr'nMlg  -\*A~*»m&**+—
 'ITTChemical Co.,  2135 Howard St., Hartville, OH  44632
  Tridom Chemical Inc., Hauttauge, N.Y.
 -NOTE:  These sources  are  not to be interpreted  as being  endorsed by  the  EPA.

-------
                                GLOSSARY

EICP - Extracted Ion Current Profile.  EICP 1s a reduction of majss
    spectrotnetrlc data acquired by continuous, repetitive measurement
    of spectra that plots the change 1n relative abundance of one or
    more Ions as a function of time.

Equivalent -  A piece of apparatus, gc liquid phase, etc. 1s
    equivalent 1f Its performance gives results Identical to or better
    than that specified.

FID -  Flame Ion1zat1on Detector

Field Blank -  Organic-free water that has been transported to the
    sampling site and bade to the analytical laboratory  in a container
    and bottle Identical to the type used to collect the samples.
    Fie^d blanks and samples must be shipped 1n separate containers.
    When received 1n the lab, the Field Blank 1s dosed,  extracted and
    concentrated as 1f .1t were an actual sample.

Limit of Detection -  10 ug/1 unless otherwise specified.  For the
    8C/MS Methods, 624 and 625, this 1s an operational Hm1t that must
    be met.  The entire analytical system must give recognizable mass
    spectra and calibration points from standards representing this
    concentration.  For Method 624, the condition (at  10 ug/1) must be
    met.when sparging 50 'ng of a component from 5.0 ml of water.  For
    Method 625, the conditions (at 10 ug/1) must be met  when Injecting
    standards at a concentration of 10 ug per ml.

Method Blank -  For Method 625, a 1-Hter volume of organic-free water
    that is extracted and concentrated to the final volume of 1 ml.
    For Method 624, a 5.0 ml aliquot of organic-free water.

Organic-Free Water -  Water that gives no interferences  when analyzed
    as a Method Blank.  See Method 624, Section 5.4, for the
    preparation of organic-free water.

Sample Blank -  Identical to Field Blank.

Significant Difference -  Check for difference in magnitude by using  a
    two-tailed Student's t-test at a 95S Confidence Level.

Significant Drift -  A 255 or larger change from an averaged
    response.  Must be based on a constantly updated  average.

-------
SIM -  Selected Ion Monitoring - Measuring the mass  spectrometrlc
    response to one or several Ions 1n real time.

Surrogate Standard -  Compounds not expected to occur  in the
    environmental sample that are added to the sample  to monitor  for
    unusual matrix effects, gross sample processing  errors, etc.   The
    compounds chosen should be chemically similar to the pollutants
    being measured.  Surrogate Standards are not used  as Internal
    standards.

Transfer P1pet -  Class A, to deliver, plpet.

-------
                Chapter IV
 QUALITY ASSURANCE AND QUALITY CONTROL PROCEDURES FOR



SCREENING AND VERIFICATION OF INDUSTRIAL EFFLUENTS FOR



                 PRIORITY POLLUTANTS
        U.  S.  ENVIRONMENTAL PROTECTION AGENCY



   ENVIRONMENTAL MONITORING AND SUPPORT LABORATORY



               CINCINNATI,  OHIO  45268
                      June  1979

-------
                                 INTRODUCTION

    "The Sampling and Analysis Procedures for Screening of Industrial
Effluents for Priority Pollutants," (revised April, 1977) was designed to
provide qualitative/semiquantitative determination of the priority
pollutants.  In order to apply these procedures to quantitative analysis,
adequate quality assurance practices must be incorporated into the
procedures.  This document provides modifications to enable implementation
of a quality assurance program that will define the accuracy and precision
of the procedures for specific parameters.

    In considering modifications to be made, the following assumptions were
made:

    1.  The program must be well documented.  It must be integrated into
        field and laboratory procedures.  It must include field and
        laboratory blanks, equipment calibration checks, replicate and
        spiked sample analyses, the analysis of check standards and
        performance evaluation standards.

    2.  The differences in industrial subcategories require method
        validation and the establishment of control parameters for each
        subcategory.

    3.  Verification Phase analyses will seldom, if ever, require analysis
        of all 114 organic priority pollutants.  Thus, the quality control
        program will focus on a smaller selected number of compounds based
        on the results of screening analyses and other data.

    4.  For purposes of the Effluent Guidelines Verfication Phase studies,
        the quality control programs provided, herein, must be applied to
        the sampling sources most frequently involved in rule making, i.e.,
        supply water, influent to treatment, and effluent from treatment.
        Primary emphasis should be placed on the effluent from treatment
        since this is where the control limits will be set.

    5.  Two distinct levels of quality control are required:  initial method
        validation and continuing performance evaluation.  The initial
        quality control is to demonstrate the method precision and accuracy
        and thus validate the method on a common waste stream (industrial
        subcategory) where waste characterization was the basis for
        establishing the subcategory.

    6.  The initial control limits will be established for each subcategory
        on the effluent from treatment prior to initiation of the routine
        sampling and analysis program for that subcategory.  The control
        limits for supply water and influent to treatment water will be
        established over a period of time using data derived from the
        ongoing continuing quality control program.

    7.  As more quality control data is obtained for each waste type within
        each subcategory, the precision and accuracy of the method will be
        updated to provide better defined control  limits.

-------
                                   FOREWORD

    These methods for sampling and analysis of industrial wastes have been
prepared by the staff of the Environmental Monitoring and Support
Laboratory, at the request of the Effluent Guidelines Division, Office of
Water and Hazardous Wastes, with the cooperation of the Environmental
Research Laboratory, Athens, Georgia.  The Quality Assurance/Quality Control
(QA/QC) procedures were prepared by the staff of the Environmental
Monitoring and Support Laboratory with the cooperation of the Effluent
Guidelines Division, Office of Water and Hazardous Wastes, the Environmental
Research Laboratory, Athens, Georgia, and several EPA Regional Laboratories.

    The procedures represent the current state-of-the-art and are applicable
to the Effluent Guidelines Survey of prirority pollutants.  This survey is
divided into two phases; the Screening Phase and the Verification Phase.

    The objectives of the Screening Phase are:

    1. To provide a determination of the presence or absence of a particular
       priority pollutant.

    2. To provide a quantitative estimate of the concentration of priority
       pollutants present.

    3. To focus the verification survey to specific priority pollutants.

    4. To provide direction to other best available technology (BAT)
       regulatory strategies.

    The objectives of the Verification Phase are:

    1. To provide the confirmation of the presence or absence of the
       priority pollutants identified during the screening phase, necessary
       for consideration in rule making.

    2. To provide qualitative pollutant data with a known precision and
       accuracy.

    3. To provide pollutant data for use in rule making.

    4. To serve as a basis for technology selection.

-------
                                  DISCLAIMER

    The mention of trade names or commercial products in this document is
for illustration purposes, and does not constitute endorsement or
recommendation for use by the U. S. Environmental Protection Agency.

-------
                                   CONTENTS

            Quality Assurance and Quality Control Procedures for
              Screening Analysis for Organic Priority Pollutants

110    Scope	 1

120    Routine Quality Assurance and Quality Control	 1

       121    Method Blank	,	1
       122    Field Blank	 2
       123    GC-MS Calibration Check	*	, 3
       124    GC Performance Check..	 4
       125    Performance Evaluation	 5

130    Reporting of Data	 5

200    Sampling Procedures for Screening Phase	 6

       201    Extractable Organics (Base Neutrals - Acids and Pesticides).. 7
       202    Preservation of Extractable Organics	 9
       203    Field Blank Procedure for Automatic Samplers	10
       204    Volatile Organics (Purgeables)	11
       205    Preservation of Volatile Organics	13
       206    Grab Samples for Volatile Organics	13


               Quality Assurance and Quality Control Procedures
           for Verification Analysis of Organic Priority Pollutants

110    Scope	14

120    Routine Quality Assurance and Quality Control	15

       121    Preliminary Clean Water Precision and Accuracy	15
       122    Method Blank	17
       123    Field Blank	18
       124    GC-MS Calibration Check	19
       125    GC Performance Check	19
       126    Performance Evaluation	20
       127    Saving Extracts	21
       128    Archiving GC/MS Data	..21

130    MethodXEvaluation	21

       131    Extractable Organics (Base Neutrals-Acids and Pesticides)....21
       132    Volatile Organics (Purgeables)	29

140    Continuing Quality Assurance and Quality Control	33

       141    Extractable Organics (First Day Samples)	33

-------
                             CONTENTS (continued)
       142    Extractable Organics (Subsequent Day Samples)	37
       143    Volatile Organics (First Day Samples)	38
       144    Volatile Organics (Subsequent Day Samples)	39
200    Sampling Procedures for Verification Phase	40
       202    Extractable Organics Base/Neutrals-Acids and Pesticides)	40
       203    Volatile Organics (Purgeables)	42
References	43
Appendices
       A  Storage of Final Extracts	....51
       B  Archiving GC/MS Data on Magnetic Tape	52
       C  Numerical Example of Verification Phase Results	55
       D  Possible Sources of Standards	58
Gl ossary	59

-------
                                    TABLES
Number
  1    DFTPP Key Ions and Ion Abundance Criteria	44
  2    BFB Key Ions and Ion Abundance Criteria	45
                                   FIGURES
Number
  1    Tailing Factor Calculation	46
  2    Summary of Initial Validation Analyses for Extractable Samples	47
  3    Summary of Initial Validation Analyses for Purgeable Samples	48
  4    Summary of Ongoing Quality Assurance for First Day Samples	49
  5    Summary of Ongoing Quality Assurance for Subsequent Day Samples	50
                                      VII

-------
         Quality Assurance and Quality Control Procedures
      for Screening Analysis for Organic Priority Pollutants

110 Scope
    This procedure is provided for use by laboratories
    performing screening analyses using the 6C/MS methods 624
    and 625.  Screening analysis provides a determination of the
    presence or absence of a specific compound and a
    semiquantitative estimate of the concentration of a specific
    compound.  Quality Assurance (QA) is the total program for
    assuring the reliability of the monitoring data.  Quality
    control (QC) is the routine application of procedures for
    controlling the measurement process.
120 Routine Quality Assurance and Quality Control
    121  Method Blank - The method blank is defined as an
         appropriate volume of "organic-free" water which has
         been processed exactly as the sample (including
         glassware, reagents, solvents, etc.).  Reagents or
         solvents having background levels that interfere with
         the compounds to be determined must be purified and
         shown to be acceptable or replaced with some that are
         acceptable prior to proceeding with analyses.  Problems
         encountered and corrective actions taken shall be
         documented and reported for the record.
         121.1  For the extractable fractions (Base/Neutral,
                Acid, and Pesticide) the method blank requires

-------
            extraction  of  1-liter  "organic-free"  water.   A
            method blank must be extracted  for  each  set  of
            field  samples  extracted  at  a given  time  (at
            least  one method blank per  20 field samples
            analyzed) and  whenever a new source of reagent
            or  solvent  is  introduced into the  analytical
            scheme.  The method blank can be screened by
            GC-FID.  Analysis by GC-MS  is required only  if
            GC-FID analysis of the field blank  gives peaks
            larger than the internal standard  peak.
     121.2   For the  volatile fraction,  5 ml of "organic-
            free"  water should be  analyzed  by  the purge  and
            trap methodology only  if positive  interferences
            are noted during the  analysis of a field blank.
            If  positive interference still  occurs, repeat
            the method  blank analysis.   If  interference
            persists, dismantle  the  system, thoroughly clean
            all parts that contact the  sample,  purge gas,
            and carrier gas.  Replace or repack the  sorbent
            trap and change purge  and carrier  gas.
122  Field  Blank
     The field  blank is defined  as an appropriate volume of
     "organic-free"  water  which  has  been to the sampling
     site and back to the  analytical laboratory in a
     container  and bottle  identical  to  the  type used to
     collect the samples.   Field  blanks and samples  must be

-------
     shipped in separate containers.   When received in the
     lab,  the field blank is dosed,  extracted and
     concentrated as if it were an  actual  sample.
     122.1   For the extractable fractions  (Base/Neutral,
            Acid, and Pesticide),  the field blank may be
            screened by 6C-FID or  6C-EC (pesticides only).
            Full  GC-MS analysis is  to be performed if the
            screening analysis gives any peaks larger than
            the internal standard  peak.
     122.2  For the volatile fraction, analyze a 5-ml field
            blank after each sample analysis.  Follow the
            guidelines in 121.2 if positive interferences
            are noted during the analysis  of a field blank.
123  GC-MS Calibration Check
     123.1  For the Base/Neutral-Acid or Pesticide fraction,
            daily analyze decafluorotriphenylphosphine
            (DFTPP) by injection into the  GC inlet.  See
            Reference 1.  (DFTPP must also be run after any
            mass spectrometer tuning is done).  This
            calibration check may  be done  in conjunction
            with Section 124.  The requirement is that 50
            nanograms of DFTPP injected must meet the
            specifications prescribed.  Table 1 lists the
            DFTPP key ions and ion abundance criteria.
     122.2  For the volatile (purgeable) fraction, analyze
            4-bromo-fluorobenzene  (BFB) daily by direct
                         osr-3

-------
            injection into the GC/MS.   The requirement is
            that 20 nanograms of PFBB  injected must meet the
            prescribed specifications.   Table 2 Itsts the
            BFB key ions and ion abundance criteria.
124  GC Performance Check
     124.1   SP-2250 column - At the beginning of each day
            that base/neutral or pesticide analyses are to
            be performed, inject 100 nanograms of benzidine
            (base neutral analyses) or  aldrin (pesticide
            analyses) either separately or as part of a
            standard mixture that may  also contain 50 ng of
            DFTPP.  Performance for base/neutrals is
            acceptable if the tailing  factor for benzidine
            is less than 3.  Performance for pesticides is
            acceptable if the tailing  factor for aldrin is
            less than 2.  Calculation  of the tailing factor
            is given in Reference 2 and described in Figure
            1.  Retain all plots of performance evaluation
            as evidence of valid performance.
     124.2  SP-1240 DA Column - At the beginning of each day
            that acid fraction analyses are to be performed,
            inject TOO nanograms of pentachlorophenol either
            separately or as part of a standard mixture that
            may also contain DFTPP.  Performance is
            acceptable if the tailing factor is less than
            5.  Calculation of the tailing factor is given
                        JSt- 4

-------
                in Reference 2 and described in Figure 1.
                Retain all  plots of performance evaluation as
                evidence of valid performance.
         124.3  Carbopak C  with Carbowax 1500 or Carbopak B with
                SP-1000 column.  At the beginning of each day
                that purgeables analyses are to be performed,
                inject 20 nanograms of BFB either separately or
                as part of  a standard mixture.   Performance is
                acceptable  if the tailing factor is less than
                2.  Calculation of the tailing factor is given
                in Reference 2 and described in Figure 1.
                Retain all  plots of performance evaluation as
                evidence of valid performance.
    125  Performance Evaluation
         Once each 20 working days; dose, extract, and analyze a
         performance evaluation sample (available from
         EMSL-Cincinnati).   Retain the results  as evidence of
         valid performance.
    126  If the final extract is to be retained, or shipped to
         another location,  follow the appropriate guidelines
         given in Appendix  A.
130 Reporting of Data
    131  All  concentrations should.be reported  in ranges, i.e.
         less than 10 ug/1, 10 to 100 ug/1, or  greater than 100
         ug/1.
    132  Report the mean and standard deviation of the relative

-------
         response ratios of the standards (10 and 100 ug/1) from
         the MS analysis with the range values in 131.
    133  Report all  quality control  data with the analytical
         results.  This includes, but is not limited to, the
         recovery of surrogate standard spikes (Method 624), the
         precision for the analysis  of surrogate standard spikes
         (Method 624), the results from the analysis of
         performance evaluation samples, and identification of
         components found in the blanks.
    134  If the 6C/MS data is to be  archived on magnetic tape,
         follow the guidelines given in Appendix B.
200 Sampling Procedures for Screening Phase
    The initial characterization (screening) of the varied
    industrial discharges covered by this program will be made
    on an analysis of a composite effluent sample.  Any scheme
    for collecting a composite sample is, in effect, a method
    for mechanically integrating to  obtain average
    characteristics of a discharge.   During the screening phase
    the sample composite can be used to determine the average
    characteristics which would be representative of that
    discharge.  Simple composite samples are those that are made
    up of a series of aliquots of constant volume collected at
    regular time intervals in a single container.  Some
    situations may require flow or time proportional sampling;
    this determination will be made by the individual project
    officer after considering his specific industrial category.
    The determination of compositing period 24, 48, or 72 hours

-------
will be made on a case-by-case basis.  The duration of
compositing will depend on the type of sample being
collected, the type of facility being sampled, and the time
varying characteristics of the discharge.  The rate of
change of flow and other characteristics of the discharge
and the accuracy required will also influence the
determination of the compositing period.  For example,
longer compositing periods would be warranted when less
stable unit process operations are being sampled.
All samples and blanks must be carefully identified using
water proof labels and water proof ink.  Include the
following information on the label:  sample number, date and
hour of sampling, complete information as to source and
sampling point, preservative added, if any, and name of
person collecting the sample (include address and/or phone
number).
201  Extractable Organics (Base/Neutrals - Acids and
     Pesticides)
     201.1  Collect a representative composite sample.  The
            maximum time interval between aliquot samples
            shall be no longer than 30 minutes.  The minimum
            aliquot size shall be TOO ml.  The sample must
            be collected with an automatic sampler using the
            equipment and methods outlined below.  Minimum
            composite volume must be 2h gallons.
     201.2  Automatic Sample Collection - A peristaltic pump
            automatic sampler with timer and a single glass
                           - 7

-------
       compositing  jug  is  required.   The 2*5-3 gallon
       compositing  bottle  must be glass and cleaned as
       outlined below.   New unused tubing must be used
       for the sampling line and for the pump for each
       individual outfall  or sample location.  Vacuum
       type automatic samplers may be used provided
       that the sample  chambers are glass and that they
       are cleaned  after every use as outlined for
       glass composite  containers.  Place the sampler
       or composite container in an insulated chest and
       ice.  Maintain the  sample at 4°C during the
       compositing  procedure.  At the completion of the
       compositing  period, seal the container with a
       Teflon lined cap.  Place the container in an
       insulated shipping  container, ice, and seal,
       then ship to the analytical laboratory.
       Maintain at  4°C during transport and storage
       prior to analysis.
201.3  When sampling raw untreated industrial
       discharges which are generally high in suspended
       solids, it is imperative that adequate sample
       flow rate be maintained throughout the sample
       train in order to effectively transport the
       solids.  In  horizontal runs, the velocity must
       exceed the scour velocity, while in vertical
       runs the settling or the fall velocity must be
       exceeded several times to  assure adequate

-------
            transport of solids in the flow.  The equipment
            used in sampling raw discharges then must have a
            minimum intake velocity of 60 cm per second (2
            fps).  In the sampling of treated effluents,
            just about any commercially available automatic
            liquid sampler could be used.
     201.4  When more than one laboratory is involved in the
            analysis of the various parameters, the sample
            should, if at all possible, not be divided in
            the field but rather at the contractors'
            laboratory.  For purpose of this program the
            composite will be divided into four parts, one
            part for metals analysis, one for pesticides and
            PCB's, one for GC/MS compounds and one for the
            classic parameters.
     201.5  Blend the composite sample to provide a
            homogeneous mixture including a representative
            suspension of any solids in the container.  No
            specific method is required; hand stirring with
            clean glass or Teflon rods, mechanical paddles
            or magnetic mixing with Teflon coated stirring
            bars may be used.  Metal mixing devices may not
            be used.
202  Preservation of Extractable Organics
     202.1  Do not add any preservative to the sample; just
            seal and prepare for shipment.  All samples must
            be carefully identified using labels supplied by
            EGD.  Indicate on the label whether the sample

                        3ST-9

-------
            is a raw discharge or treated effluent as
            shown.  If sample is to be run on the plasma
            unit, only indicate so at base of tag.
203  Field Blank Procedure for Automatic Samplers
     203.1  Blank Water - Blank water must be as free from
            organic interferences as possible.  The
            analytical laboratory should supply this water
            in bulk glass containers (minimum of five
            liters) for field use.  The supplying laboratory
            shall analyze the blank water to determine the
            organic background that may be present.
     203.2  Procedure - All parts of the sampling system
            must be scrubbed with hot detergent water and
            thoroughly rinsed with tap water and blank water
            prior to use.  Further rinsing with methylene
            chloride is required when parts permit, i.e.,
            are not susceptible to dissolution by the
            solvent.  (Note:  Tygon plastic tubing is a
            source of phthalate ester contamination.  Where
            its use is required, i.e., in the peristaltic
            pump, the length must be kept as short as
            possible.)  Teflon  is acceptable and may be used
            in other parts of the sampling system as in
            intake lines.  In the field, pump two liters of
            blank water through the sampling line and pump
            tubing and discard.  Then pump three  liters of
                           "- 10

-------
            blank water through the system and collect as a
            blank in a 1-gallon sample bottle that has been
            prepared as described below.   Seal the bottle
            with a Teflon lined cap.  Immediately ice the
            blank (4°C) and maintain at 4°C during the
            transport and storage prior to analysis.
     203.3  Composite Container - Prepare narrow-mouth glass
            sample bottles for use by washing with hot
            detergent water and thoroughly rinsing with tap
            water and blank water.  Heat the bottles at
            400°C in a muffle-furnace or dry heat
            sterilizer for 30 minutes or alternatively;
            rinse with methylene chloride and air dry at
            room temperature protected from atmospheric or
            other sources of contamination.  Caps for the
            bottles must be lined with Teflon which has been
            solvent rinsed as above.
204  Volatile Orgam'cs (Purgeables)
     204.1  Containers - Use 45 to 125 ml screw cap glass
            vials with Teflon faced silicone septa:
            204.1.1  Vials - Pierce #13074 or equivalent.
            204.1.2  Septa - Pierce #12722 or equivalent.
     204.2  Wash the bottles, septa, and caps with hot water
            and thoroughly rinse with tap water and blank
            water.  Heat the bottles and septa at 105°C
            for one hour, cool to room temperature in an

-------
       enclosed contaminant-free area.  When cool, seal
       bottles with septa (Teflon side down) and screw
       cap.  Maintain the bottles in this condition
       until just prior to filling with blank water or
       sample.
204.3  Collect duplicates 45-125 ml samples each time
       samples are collected.  Two blank water samples,
       sealed in 45 ml vials, are to accompany the
       sample bottles during shipment to and from the
       sampling site.  If preservation for residual
       chlorine is to be used, collect four samples
       during each sampling period.  Two should be
       preserved and two not preserved.  Two preserved
       and two non-preserved blanks are to be provided.
204.4  Filling and Sealing Bottles - Slowly fill each
       container to overflowing.  Carefully set the
       container on a level surface.  Place the septum
       (Teflon side down) on the convex sample
       meniscus.  Seal the sample with the screw cap.
       To ensure that the sample has been properly
       sealed, invert the sample and lightly tap the
       lid on a solid surface.  The absence of
       entrapped air bubbles indicates a proper seal.
       If air bubbles are present, open the bottle, add

-------
            additional sample, and reseal.  The sample must
            remain hermetically sealed until it is analyzed.
205  Preservation of Volatile Organics
     205.1  Preservative (sodium thiosulfate or sodium
            bisulfite) is used to stabilize samples
            containing residual chlorine.  The production of
            chloroform and other haloforms continues in such
            samples if they are not stabilized.  Waste
            streams that have been treated with chlorine
            should be tested on-site to determine whether or
            not preservative is needed.  If preservation is
            required, collect both preserved and non-
            preserved samples.  Wrap the samples with water
            proof packing material, place in an insulated
            chest and ice at 4°C.  Maintain at 4°C
            during transport and storage prior to analysis.
206  Grab Samples for Volatile Organics Q^rgeables)
     206.1  Collect grab samples (minimum of one per day)
            for the analysis of volatile organics.  Collect
            samples from the raw process discharge, the
            treated effluent, and the treated effluent after
            chlorination, when chlorination is practiced.
            It is recommended that the samples be collected
            from mid-channel at mid-depth.  Samples should
            be collected at a turbulent, well mixed section
            of the channel.

-------
             Quality Assurance and Quality Control Procedures for
             Verification Analysis For Organic Priority Pollutants
110 Scope
    111   These procedures are provided for use by laboratories performing
         verification analyses using EPA Methods 624 and 625.  Verification
         analyses require quantitative determinations with known precision
         and accuracy, so that these data will be acceptable for use in
         developing effluent limitation guidelines.   To provide data with
         this degree of reliability, a strong quality assurance and quality
         control  program is presented.  These procedures are designed to
         produce  data with known precision and accuracy so that a
         determination of confidence can be placed in the data.  Quality
         Assurance (QA) is the total program for assuring the reliability of
         the monitoring data.  Quality control (QC)  is the routine
         application of procedures for controlling the measurement process.
    112  Initially, the methodology must be validated for each industrial
         subcategory.  The requirement for validation of each subcategory is
         based on the unique nature of the wastewater associated with most
         of the subcategories.  Since the effluent from treatment is to be
         used for setting control limits, it will be used to develop initial
         validation data for the method prior to routine sampling and
         analyses.  Based on screening data, a particular subcategory may
         not require verification analyses of all four fractions (Volatiles,
         Base/Neutrals, Acids, and Pesticides) or for all of the individual
         Consent  Decree compounds, in which case, the method requires
         validation only for the fraction or the selected compounds of
         interest.
                                       '-14

-------
    113  The results of the validation analyses will  be used to provide
         information with which to judge a laboratory's ability to interpret
         and implement the method for each industrial subcategory.  Intial
         QC limits  for precision and accuracy will  be established using
         these results, and then used in subsequent analyses as control
         limits.  A numerical  example is given in Appendix C.
    114  After the  method is validated for each subcategory and routine
         analysis begins, continuing QA/QC will be required to ensure that
         the subsequent analyses are within the established control limits.
    115  Prior to developing initial method validation and a continuing
         quality  control program, the analyst (individual or group if team
         approach is used) must demonstrate the ability to perform the
         required analyses.  If a laboratory has not  established precision
         and accuracy criteria for clean water, the laboratory must perform
         replicate  analyses of clean water as prescribed in section 121.
         Such data  must be presented to the responsible EGD project officer
         prior to the beginning of the verification phase.
120 Routine Quality Assurance and Quality Control
    121  Preliminary Clean Water Precision and Accuracy
         121.1  Before any work is begun on actual  field samples, a labora-
                tory must demonstrate its ability to  properly perform the
                liquid-liquid extractions, the gas purge extractions, and
                the required chromatography.  Clean water spikes are ana-
                lyzed to demonstrate the laboratory's ability to implement
                                     -15

-------
       Methods 624 and 625, and to establish the baseline precision
       and accuracy criteria for the method in that laboratory'.
121.2  Procedure:
       121.2.1  Prepare "organic-free" water for use in determining
                preliminary precision and accuracy according to the
                procedures given in Methods 624 and 625.
       121.2.2  Spike four replicates of clean water with each
                compound of interest at a concentration
                approximately equal to 10 times the limit of
                detection.  In addition, spike all purgeable
                aliquots with a minimum of three surrogate
                standards at a level of 100 ug/1.   For extractable
                organics, each replicate must be one liter; each
                purgeable sample requires at least 100 ml.  Do not
                dose purgeables with more than 20 ul of an
                alcoholic standard per each 100 ml of water.
                Analyze spiked solutions according to Method 624 or
                625.
121.3  Precision - For each parameter, use the resulting observed
       values of the spikes (0^ 02, 03, and 04) to
       calculate the standard deviation (S) of the replicates
       according to Equation 1.

                           n(n-l)

                n = number of replicates
                            -16

-------
     121.4  Accuracy - For each parameter, use the resulting observed
            values of the spikes (0-p C^, Og, and 04) to
            calculate the mean percent recovery (P) of the method
            according to equation 2.
Eq. 2
Where:
P = 100 (£ 0-j )
nT
                     n a number of replicates
                     T = true value of the spike
     121.5  The precision and accuracy data shall be documented for the
            record as evidence that the laboratory can properly perform
            the extractions and chromatography essential for methods 624
            and 625.
122  Method Blank - The method blank is defined as an appropriate volume
     of "organic-free" water which has been processed exactly as the
     sample (including glassware, reagents, solvents, etc.).  Reagents
     or solvents having background levels that interfere with the
     compounds to be determined must be purified and shown to be
     acceptable or replaced with some that are acceptable prior to
     proceeding with analyses.  Problems encountered and corrective
     actions taken shall be documented and reported for the record.
     122.1  For the extractable fractions (Base/Neutral, Acid, and
            Pesticide) the method blank requires extraction of 1-liter
            "organic-free" water.  A method blank must be extracted for
            each set of field samples extracted at a given time (at
            least one method blank per 20 field samples analyzed) and
                              lt-17

-------
            whenever  a new source  of reagent or solvent is introduced
            into  the  analytical  scheme.   The method blank can be
            screened  by GC-FIC.  Analysis by 6C-MS is required only if
            GC-FID analysis of the field blank  gives any peaks larger
            than  the  internal  standard peak.
     122.2  For the volatile fraction, 5 ml  of  "organic-free" water
            should be analyzed by  the purge  and trap methodology only if
            positive  interferences are noted during the analysis of a
            field blank.  If positive interference still occurs, repeat
            the method blank analysis.  If interference persists,
            dismantle the system,  thoroughly clean all parts that
            contact the sample,  purge gas and carrier gas.  Replace or
            repack the sorbent trap and change purge and carrier gas.
123       Field Blank - The field  blank is defined as an appropriate
          volume  of "organic-free" water which has been to the sampling
          site and back to the analytical laboratory in a container and
          bottle  identical to the type used to collect the samples.
          Field blanks and samples must be shipped in separate
          containers.  When received in the lab, the field blank is
          dosed,  extracted and concentrated as if it were an actual
          sample.
     123.1  For the extractable factors  (Base/Neutral, Acid, and
            Pesticide), the field blank may be screened by GC-FID or
            GC-EC (Pesticides) only).  Full  GC-MS analysis is to be
            performed if the screening analysis gives any peaks  larger
            than the  internal standard peak.
                              ISf-18

-------
     123.2  For the volatile fraction, analyze a 5-ml field blank after
            each sample analysis.  Follow the guidelines in 121.2 if
            positive interferences are noted during the analysis of a
            field blank.
124  GC-MS Calibration Check
     124.1  For the Base/Neutral-Acid or Pesticide fraction, daily
            analyze decafluorotriphenylphosphine (DFTPP) by injection
            into the GC inlet.  See Reference 1.  (DFTPP must a-lso be
            run after any mass spectrometer tuning is done).  This
            calibration check may be done in conjunction with Section
            125.  The requirement is that 50 nanograms of DFTPP injected
            must meet the specifications prescribed.  Table 1 lists the
            DFTPP key ions and ion abundance criteria.
     124.2  For the volatile (purgeable) fraction, analyze
            4-bromo-fluorobenzene (BFB) daily by direct injection into
            the GC/MS.  The requirement is that 20 nanograms of BFB
            injected must meet the prescribed specifications.  Table 2
            lists the BFB key ions and ion abundance criteria.  This
            calibration check may be done in conjunction with Section
            125.
125  GC Performance Check
     125.1  SP-2250 column - At the beginning of each day that
            base/neutral or pesticide fraction analyses are to be
            performed, inject 100 nanograms of benzidine (for
            base/neutral analyses) or aldrin (pesticide analyses) either
            separately or as part of a standard mixture that may also
            contain 50 ng of DFTPP.  Performance for base neutrals

-------
            is  acceptable  if  the  tailing  factor  for  benzidine is less
            than  3.   Performance  for  the  pesticides  is  acceptable if the
            tailing  factor  for  aldrin is  less  than 2.   Calculation of
            the tailing factor  is given in  Reference 2  and  described in
            Figure 1.  Retain all plots of  performance  evaluation as
            evidence of valid performance.
     125.2   SP-1240  DA Column - At the beginning of  each  day that acid
            fraction analyses are to  be performed, inject 100 nanograms
            of pentachlorophenol  either separately or as  part of a
            standard mixture  that may also  contain DFTPP.  Performance
            is  acceptable  if  the  tailing  factor  is less than 5.
            Calculation of  the  tailing factor  is given  in Reference 2
            and described  in  Figure 1. Retain all plots  of performance
            evaluation as  evidence of valid performance.
     125.3   Carbopak C with Carbowax  1500 or Carbopak B with SP-1000
            column.   At the beginning of  each  day that  purgeables
            analysis are  to be  performed, inject 20  nanograms of BFB
            either separately or  as part  of a  standard  mixture.
            Performance  is  acceptable if  the tailing factor is less than
            2.  Calculation of  the tailing  factor is given  in Reference
            2 and described in  Figure 1.  Retain all plots  of
            performance evaluation as evidence of valid performance.
126  Performance  Evaluation - Once each 20  working days;  dose, extract,
     and analyze  a performance  evaluation sample (available from
     EMSL-Cincinnati).   Retain  the results  as  evidence  of valid
     performance.
                                   20

-------
    127  Saving Extracts -If the final extract is to be retained, or shipped
         to another location,  follow the appropriate guidelines given in
         Appendix A.
    128  Archiving GC/MS Data - If the 6C/MS data are to be archived on
         magnetic tape, follow the guidelines in Appendix B.
130 Method Validation
    131  Extractable Organics fBase/Neutrals-Acids and Pesticides) - The
         following procedures are to be applied, separately, to samples
         being analyzed for the Base/Neutral-Acid group of compounds and to
         samples being analyzed for the Pesticides group.  The analyses
         shall be performed according to the procedures given in Method
         625.  The validation studies must be performed under the same
         conditions ordinarily applied to the samples of a given
         subcategory.  That is, if separatory funnels are routinely used for
         extraction of the samples, the study must be conducted using
         separatory funnels.  If continuous extractors are used for routine
         analysis of the subcategory, the validation study must be performed
         using the same type of continuous extractors.
         131.1  Sample pretreatment - The laboratory will receive a 24-hour
                composite sample of adequate volume to carry out the
                validation study and one field blank taken as described in
                the sampling protocol on the same day from the same source.
                Mix composite sample and withdraw a 1-liter aliquot for
                analysis.  Vigorously mix the sample by some type of
                stirring device.  Withdraw aliquots, while stirring, into a
                1-liter graduated cylinder, using a siphon made of glass or
                                     - 21

-------
       Teflon.   Measure and  record the volume.   Transfer the
       aliquot  to  a 2-liter  separatory funnel  or continuous
       extractor for spiking.   Initially analyze a 1-liter aliquot
       of the sample to determine the sample background so that
       proper spiking levels can be selected for 131.2.1.   The
       remainder of the composite sample should be stored  at 4°C
       until  the validation  study is begun.  At the same time,
       analyze  a 1-liter aliquot of the field  blank.   Choose three
       levels of priority pollutant spikes to  cover the expected
       concentration range of the samples in the subcategory.
131.2  Preparation of Aliquots for Validation  Study - Withdraw
       twelve 1-liter aliquots from the stirred composite  sample as
       described in 131.1.  Separate into three groups of  four.
       131.2.1   Spiking of Aliquots - Spike two aliquots of each
                group with surrogate standards only.   The  other two
                aliquots are spiked with surrogate standards plus
                the priority pollutant standards of interest at one
                of the concentration levels (See Figure 2).  Repeat
                this process for each group of aliquots.  Select
                the three spiking concentrations for the priority
                pollutants based on the results of the background
                analysis obtained in 131.1.  If the initial
                background level for a particular priority
                pollutant is X, select the three spiking levels to
                give final concentrations of 2X, 10X, and 100X.  If
                X  equals 15 ug/1, dose with 15, 135, and 1485 ug
                        -DT-22

-------
         per liter.   This gives final concentrations of
         priority pollutants of 30, 150, and 1500 ug/1.
         Spike each  1-liter replicate with each surrogate at
         a level of  100 ug/1.  (Note:  Consideration should
         be given to the water solubility of the compounds
         being spiked when selecting the spiking
         concentration levels.)
131.2.2  Prepare spiking standards in concentrations such
         that no more than 5 ml of spiking solution is added
         for each liter of sample.  This will ensure that
         the solubility of the standard in water will not be
         significantly affected by the added organic
         solvent. Add the spiking solution to the sample
         aliquots in the separatory funnel using a transfer
         pipet.  After adding the spikes, thoroughly mix the
         samples and after one hour at room temperature
         proceed with the extraction.  If continuous
         extractors  are used, it may be necessary to spike
         the 1 liter aliquots before they are placed into
         the extractor.  Place the aliquot in a separatory
         funnel or a clean bottle, spike, and transfer to
         the extractor. . The bottle must be rinsed with
         solvent.  Wait an hour and begin the extraction.
         (Note:  Extractions of all samples must begin
         within 48 hours of receipt in the laboratory.)

-------
131.3  Use of the Data from Spiked Samples in Analyses - The data
       obtained from the determination of priority pollutants are
       used to calculate the precision and accuracy of the method
       and to establish control  limits for the individual compounds
       of interest.   Surrogate spikes are added to every sample to
       provide quality control on every sample by monitoring for
       matrix effects and gross sample processing errors.  The
       surrogate is  not used as an internal  standard for
       quantification purposes.   Suggested surrogate spikes are
       given in Section 6.4 of Method 625.  If verification is
       needed for only one fraction, only the surrogates for that
       fraction shall be added.
131.4  Extract and analyse all aliquots as directed in Method 625
       or other appropriate EPA methods„
131.5  Calculation of Precision and Accuracy - The precision of the
       method may be calculated from the data obtained during the
       validation study.  There are three spiked concentration
       levels of priority pollutants as outlined in Figure 2. The
       method precision for the background level of priority
       pollutants occurring in the sample may be calculated from
       the three pairs of replicate aliquots which are spiked only
       with surrogate standards (Al, A2; Bl, B2; Cl, and C2.  See
       Figure 2).  The precision and accuracy for the surrogates
       may be calculated from all twelve replicates since the
       spiking level is constant for both sets of six samples.  The
       precision and accuracy for the surrogates may be calculated

-------
for each set of six samples  if there  is  an  effect  due  to  the
added priority pollutant spike.  Similarly, the precision
and accuracy at each spike level of the  priority pollutants
may be determined from the two replicate aliquots  that
received that spike (Dl, D2; El, E2;  Fl, and F2 See Figure
2).
a.  Precision
Calculate the range (R) for  each pair of replicate aliquots
ie., duplicate analyses, according to equation 3:
      Eq. 3   R =|x] - X2|
Where:
     X, and %2 are eacn an analytical result from  two
     replicate aliquots.
The concentration level related to R  can be represented as
in equation 4.

       Eq. 4    I =  (X1 + X2)
                         2
Where:
        X is the mean of the duplicate analyses, X, and
        V
For any group of n duplicate analyses that  are considered
similar to each other, their ranges (Rj and means (X.).
where i = 1 to n, can be used to estimate the critical
difference (Rc) between similar future duplicate analyses
or any specific concentration level (C).  Calculate R  as
shown in equation 5.
                  37- 25

-------
    Eq.  5      R  -  3.27(C)
                         n
From these data develop a table of R  values for various
                                    C
concentration (C) values that span the concentration range of
interest.  Use these Initial critical difference values to
judge the acceptability of succeeding duplicate results
generated under the same conditions.  Revise and update as
additional duplicate data becomes available.  When more than.
15 pairs of duplicates are available within any specific
concentration level C, R  should be calculated directly
from the average range of these duplicates alone.  Equation  5
reduces to

                R  =  3.27(C)    £  R-J
                         n      (.1=1
   b.   Accuracy for Priority Pollutants
        Calculate the background concentration of
        priority pollutants occurring in each of the
        field composite samples.  The calculation  is
        similar to equation 4, but there are six
        pieces of data (Al, A2, Bl, 82, Cl, C2)
        available for this calculation as shown in
        Figure 2.  Therefore, the calculation is as
        shown in equation 6.
                             6
          Eq. 6      *=  (  £ X.)
                      "-  26

-------
 Where:
       «•
       X is the mean X., i-1 to 6 are the analytical
       results for the six 1-liter aliquots of a single
       composite sample spiked with surrogates only.
       Calculate the recovery of each priority pollutant
       in each of the 1-liter aliquots spiked with
       priority pollutants (01, D2. El, E2, Fl and F2)
       according to equation 7:

           Eq. 7    P » 100 (Z - I)
                           T

Where:
       P is the percent recovery of the spike
       Z is the analytically determined
       concentration of the priority pollutant in
       the spiked aliquot
       X is the mean background concentration of
       the priority pollutant and
       T is the true value of the spike.

Determine the percent recoveries for each priority
pollutant at all of its concentration levels.  If
there is no significant difference between the
percent recoveries for the various concentration
               J7-27

-------
   levels, all n of the percent recovery values may be
   treated together as in equations 8 and 9.  If some
   of the percent recovery values are significantly
   different, each group of similar percent recover ie.s
   must be treated independently to develop its own
   characteristic mean percent recovery (P) and its
   associated standard deviaion (S ).
                n
    Eq. 8  "P *  £   Pi
Where:
    P is the mean percent recovery
    P is an individual percent recovery value
    n is the number of observations at this
      concentraton level
                      '28

-------
                                n(n-l)
            c.  Accuracy for Surrogates
            Proceed exactly as with priority  pollutants  in  131.5b  above,
            keeping the following two differences  in mind:  there  is  no
            background concentration and there are six sets of  duplicate
            analyses for the surrogate spikes; three sets spiked with
            priority pollutant (Dl, D2; El, E2; Fl, F2)  and three  sets
            without (Al, A2; Bl, B2; Cl, C2), see  Figure 2.  Calculate
            the percent recovery as shown  in  equation 10.

                      Eq. 10  P = 100 Z

                 Where:
                       P is the percent recovery of the surrogate  spike.
                       Z is the measured value of the surrogate spikes
                       in the aliquot.
                       T is the calculated or true value of the
                       surrogate spikes added to the sample.
            Calculate the mean percent recovery (P) and the standard
            deviation (S )  of the percent recovery of the surrogate
            spikes in all  of the sample aliquots according to equations
            8 and 9.
132  Volatile Organics (Purgeables)
     132.1  All  individual  grab samples collected during the specified
            time period shall  be composited in the laboratory,  spiked
            and  analyzed as soon  as  possible.   Significant losses of
                             17  29

-------
       some of the purgeable priority pollutants will  occur in 1 to
       2 days.   However,  in  a nonreactive matrix,  most are stable
       up to 14 days.   The method validation study requires twelve
       5-ml  aliquots  from each composited sample.   Carefully
       transfer the remaining volume of composited sample back to a
       clean vial  or vials and seal  with no headspace  as done when
       collecting a sample.   These should be held at 4°C until it
       is determined  that there is no further need for the sample.
       Figure 3 summarizes the validation study for volatile
       organ ics.
       Caution:  Prepare  only as many sample aliquots  as can be
       analyzed in the working day.   This may mean that each of the
       three concentration levels will be analyzed on  different
       days.
132.2  Pretreatment of Grab  Samples - Individual grab  samples
       should be composited  according to the following procedure:
       a.   Composite only grab samples of equal volume.
       b.   Carefully pour the contents of all individual grab
            samples collected from a given source during the
            specified time period into a 1000-ml round-bottom flask
            which is chilled in a wet ice bath.
       c.   Stir the mixture gently with a glass rod for
            approximately one minute while in the ice  bath.
       d.   Carefully fill 13 clean 40-ml vials or three 120-ml
            vials and four 40-ml vials with composited sample.
       e.   Take one 40-ml vial for immediate analysis to determine
            the background of the purgeable priority pollutants.
                             30

-------
       f.   Store the remaining vials at 4°C until the validation
            study is begun.
132.3  Spiking levels for priority pollutants and surrogate
       standards - The spiking levels of the priority pollutants
       are determined by the background (X) in the samples.  The
       low level spike will give a final concentration that is 2
       times the background level.  The intermediate and high level
       spikes will give final concentrations that are 10 and 100
       times the background level.  Concentrations in excess of
       1000 ug/1 are likely to flood the gas chromatographic
       column.  Therefore, the total concentration (background plus
       spike) of each individual priority pollutant should not
       exceed 900 ug/1.  Even at this level the solubility of the
       compounds in the sample must be considered.  The spiking
       level for all surrogate standards should be 100 ug/1.
132.4  Spiking Procedures
       132.4.1  Preparation of Spiking Standards - Prepare
                methanolic stock standard solutions of the priority
                pollutants and the surrogate standards according to
                the directions given in Method 624.
                From the methanolic primary dilutions prepare
                secondary aqueous spiking mixtures of the surrogate
                standards so that 20.0 ul of the primary standard
                solution, diluted to 50.0 ml in organic free water
                will permit adding 5 ul of the resulting solution
                to the 5 ml sample giving the desired surrogate
                            31

-------
         concentration level of 100 ug/1.
         Prepare spiking mixtures of the priority pollutant
         standards in methanol so that 20.0 ul of the
         solution added to 100.0 ml of sample will give the
         desired concentration levels.
132.4.2  To minimize the solubility effect of methanol on
         the constituents to be measured, do not inject more
         than 20 u.l of spiking solution per 100 ml of
         sample.  Never use a pipet to transfer samples or
         aqueous standards that are to be analyzed for
         volatile purgeable compounds.  Transfer samples by
         pouring into the receiving vessel.
132.4.3  Spiking the Sample Aliquots - Take one of the
         120-ml or 3 of the 40 ml sample aliquots from cold
         storage, equilibrate to room temperature, and fill
         a 100 ml volumetric flask to mark with the sample.
         Rapidly inject 20 ul of the methanolic solution of
         priority pollutant spiking standard (concentration
         2X) into the expanded area of the flask below the
         neck.  Stopper and mix by gently inverting the
         flask three times.  Fill two 5-ml syringes with
         spiked sample from the flask as directed in the
         analytical protocol.  Open the valve of the syringe
         and inject 5 ul of the surrogate standard spiking
         solution.  Inject the sample aliquot into the
         purging device and analyze according to Method 624.
                  '4?  32

-------
                         Take one of the 40-ml sample aliquots from cold
                         storage, equilibrate to room temperature and fill
                         two 5-ml syringes with the sample as directed in
                         Method 624.  Spike 5 ul of the surrogate standard
                         water solution (concentration 100 ug/1) into the
                         syringe through the valve giving a final
                         concentration of 100 ug/1.  Inject the sample
                         aliquot into the purge device and analyze according
                         to Method 624.  See Figure 3.  Repeat this
                         procedure twice, giving three sets of analyses of
                         two samples spiked with surrogate standards only
                         and two samples spiked with surrogate standards and
                         priority pollutants.
         132.5  Calculation of Precision and Accuracy - The precision and
                accuracy for the purgeable priority pollutants and the
                surrogate standards are calculated as directed for the
                semivolatile solvent extractable compounds in paragraphs
                131.5a,  b, and c.
140 Continuing Quality Assurance and Quality Control
    141   Extractable  Organics  (First Day Samples) - The following
         procedures should be applied to the first day samples of the
         pesticides group Base/Neutral-Acid group.  They should be carried
         out on each  first day composite sample for each new subcategory.
         This includes samples of the influent to treatment, the effluent
         from treatment  and the supply water.  An outline diagram for first
                                  'IT-33

-------
day ongoing qualityassurance samples is given in Figure 4.
141.1  Withdraw three 1-liter aliquots of the composite sample-
       according to the procedure in 131.1.
141.2  Spiking the Sample Aliquots - Spike one of the aliquots with
       priority pollutant standards plus the surrogate standards
       and two of the aliquots with surrogate standards only.
141.3  Add a spike sufficient to approximately double the
       background concentration of the priority pollutants as
       determined in 131.5b.  If the original concentration is
       higher than the midpoint of the calibration curve, then the
       concentration of the spike should be approximately one-half
       the original concentration.  Surrogate spikes as specified
       in 131.3 should be added to all three aliquots from each
       sample at a concentration level of 100 ug/1.
141.4  Analyze according to Method 625.
141.5  Calculations of Precision and Accuracy
       a.   For the first day samples, calculate the precision of
            the duplicate analyses (X, and X-) from the two
            1-liter aliquots for the priority pollutants background
            and the surrogate standards.  Calculate the range  (R)
            of the results according to equation 11.
                  Eq. 11   R =
Xl -
            The concentration of each compound is represented by
            the mean of the duplicate values.  Calculate the mean
            (X) according to equation 12.
                   Eq. 12   X = (Xi + X?)
                            - 34

-------
   Refer to the table of critical range values developed in
   131. 5a, to find the concentration (C) nearest to X.
   Use this R  to evaluate the acceptability of R from
   Eq. 11.  If R is greater than R , the system precision
   is out of control and the source of this unusual
   variability should be identified and resolved before
   continuing with routine analyses.  After correcting the
   source of this unusual variability, reanalyze the sample
   if possible.  Record the results of all duplicate
   analyses and periodically (after 5-10 additional
   duplicate results are obtained), revise, update, and
   improve the table of critical range values.
b. Accuracy for Surrogate Spikes
   Calculate the recovery of the surrogate spikes in the
   duplicates according to equation 13.

            Eq. 13    P - 100 Z

   Where :
     P is the percent recovery.
     Z is the analytically determined concentration of the
     surrogate standard spikes.
     T is the true value of the surrogate standard spikes
     added in 132.4.3.
   If the percent recoveries are not within the interval P
   +_ 3S  as determined in 131. 5c, the system should be  •
   checked for problems.  If problems exist, they must be
                     ' 35

-------
    resolved before continuing with routine analyses.
    Record the recovery of all surrogate spikes and
    periodically (every 5 to 10 additional data points),
    revise, update and improve the recovery criteria.
c.  Accuracy for Priority Pollutant Spikes
    Using the results obtained from the 1-liter aliquot of
    composite sample spiked with surrogate standards and
    priority pollutant standards, calculate the recoveries
    of the priority pollutants according to equation 14.

                Eq. 14   P = 100 (Z-X)
                                 T
   v Where:
          P is the percent recovery
          Z is the analytically determined
          concentration of the priority pollutant
          spikes
          T is the true value of the priority
          pollutant spikes added in 132.4.3, and
          X is the mean concentration of the priority
          pollutant background determined by equation
          9.
    If the percent recovery is not within the  interval
    of P _+ 3S , as determined in 131.5b the system
    should be checked for problems.  If problems
    exist, they must be resolved before continuing
                   -OT  36

-------
                with routine analyses.  Record the recovery of all
                spikes and periodically revise, update, and
                improve the accuracy criteria.
142  Extractable Organics - (Subsequent Day Samples) - The
     following procedures should be applied to each subsequent day
     sample of the Base/Neutrals-Acids group and the pesticide
     group.  A flow diagram for each subsequent day ongoing
     quality assurance samples is given in Figure 5.
     142.1  Withdraw a one-liter aliquot as directed in 131.1
     142.2  Spike the aliquot with surrogate standards at a
            concentration of 100 ug/1.
     142.3  Analyze according to Method 625.
     142.4  Determine the percent recovery of the surrogate
            standards using Equation 10.  If the percent recovery
            is outside the interval P + 3S  as determined in
            131.5c, the analytical system should be checked for
            problems.  If problems exist, they must be resolved
            before continuing further sample analyses.
     142.5  A field blank for the day must be analyzed according
            to Method 625.  If priority pollutants are found and
            quantified, the values for the field blank should be
            noted and reported along with sample results.  If
            significant interference problems occur, the method
            blank must be analyzed to determine if interference
            was introduced in the
                              "5SL  37

-------
               field or the laboratory.  Appropriate action must
               be taken to eliminate the problem before continuing
               with the analysis of routine samples.
143  Volatile Organics (First Day Samples) - Continuing quality
     assurance for the purgeable volatile organics should be
     applied to samples from influent to treatment, effluent from
     treatment, and supply water.  The following procedures should
     be carried out on each first day composite samples from the
     three sample sources from each subcategory.  An outline is
     given in Figure 4.
     143.1  Composite all relevant grab samples as described in
            Section 132.2.  Prepare six 5-ml aliquots for analysis.
     143.2  Spike two aliquots with the priority pollutant
            standards at a level twice that determined in Section
            132.5 and the surrogate standards using the procedures
            in Section 132.3 and 132.4.  Spike four 5-ml aliquots
            with surrogate standards only as in 132.3 and 132.4.
     143.3  Analyze one of the duplicates spiked with priority
            pollutants and surrogate standards and two of the four
            replicates spiked with surrogate standards only.  The
            remaining spiked aliquots  are analyzed only if  a
            problem is encountered with the analysis of the first
            set  of  aliquots.
                                 f-38

-------
     143.4  Analyze the spiked aliquots according to Method 624.
     143.5  Calculate the precision and accuracy as directed for
            the semivolatile solvent extractables as directed in
            141.5.
144  Volatile Organics (Subsequent Days Samples) - The following
     procedures should be applied to each subsequent day sample of
     the volatile organics group.  An outline is given in Figure 5.
     144.1  Composite the relevant grab samples as described in
            Section 132.2.  Prepare two 5-ml aliquots for analysis.
     144.2  Spike both aliquots with surrogate standards only to
            give a concentration of 100 ug/1.
     144.3  Analyze one of the aliquots according to Method 624.
            The other aliquot is analyzed only if a problem is
            encountered.
     144.4  Determine the percent recovery of the surrogate
            standards using Equation 10.  If the percent recovery
            fs outside the interval P + 3 S_ as determined in
                                      ~    P
            131.5c, then the analytical system should be checked
            for problems.  If problems exist, they must be
            resolved before continuing further sample analysis.
     144.5  Analyze a field blank representing the same day that
            the samples were collected.  Follow the guidelines
            given in 142.5.
                                  39

-------
200 Sampling Procedures for Verification Phase
    201  These procedures are given as an example for the collection
         of samples.
    202  Extractable Organics (Base/Neutrals-Acids and Pesticides)
         202.1  Initial QA/QC - Three 2*s-galien samples must be taken
                concurrently over the initial period of sampling (24
                hrs.) effluent from treatment if both the
                Base/Neutrals-Acid group and the Pesticide group are
                to be determined.  If only one of the two groups is to
                be determined, collect two 2H-gal1on samples or one
                5-gallon sample.  The samples can be collected using
                two or more automatic Isco samplers that have the
                sample lines linked together to ensure close replicate
                samples.
                   a.   Cluster the inlet tubes to ensure that samples
                        are taken geographically close.
                   b.   Link the compositors in a master/slave
                        arrangement, wherein one "master" sampler
                        supplies the sample  interval timing for the
                        "slave" sampler(s).  The Master sampler
                        operates in a TIME mode and the interval
                        between samples is set on the SAMPLE RATE
                        switch  (Model 1580)  or the SAMPLE INTERVAL
                        switch  (Model 1680).  The "slave" sampler(s)
                        operate in a FLOW mode, with the FLOW
                        MULTIPLIER switch (Model 1580) set on 1, or
                                    i.r-40

-------
            the SAMPLE INTERVAL switch (Model 1689) set to 1 FLOW
            PULSE.  When the master sampler times out and collects
            a sample, a flow pulse is sent to the slave
            sampler(s), causing them to collect samples
            simultaneously with the master sampler.  (Note that
            depressing the MANUAL ADVANCE pushbutton on the master
            sampler will cause the slave sampler(s) to collect a
            sample; therefore, the slave sampler(s) should be left
            in STANDBY until the master sampler set-up has been
            completed).  No modification to the equipment is
            necessary.  Use of two "slave" samplers require a "Y"
            connect cable assembly for the master samples and two
            sampler-flow meter cables to connect to the slave
            samplers'  .  Samplers should be checked before use
            in the field to ensure that each collects the required
            volume of sample +_ 5%.
     202.2  Continuing QA/QC - One 2%-gallon sample is required
            for the continuing QA/QC work.
Available from Isco, Environmental Division, Building
1020, Lincoln Air Park West, Lincoln, Nebraska, 68524.
                                   41

-------
     202.3  Blanks - One 2-gallon aliquot is needed for the preliminary
            QA/QC.  This should be obtained by flushing the organic-free
            water through all Isco samplers used.  One 1-gallon blank is
            needed for continuing QA/QC work.
203  Volatile Organics (Purqeables)
     203.1  Sampling Precautions
            a. Collect 2 replicates from each sample source (i.e.,
               duplicate samples).
            b. Fill the sample bottles in such a manner that no air
               bubbles pass through the sample as the bottle is being
               filled.
            c. Seal the bottles so that no air bubbles are entrapped in
               them.
            d. Maintain the hermetic seal on the sample bottle and keep
               at 4°C until time of analysis.
     203.2  Initial QA/QC - A minimum of 1000 ml (twenty-five 40-ml
            vials) must be collected over the relevant time period at
            each sample point.
     203.3  Continuing QA/QC - A minimum of 200 ml (five 40-ml vials)
            must be collected over the relevant time period at each
            sample point.
     203.4  Blanks - Three 40-ml "trip" blanks (field blanks) are
            required for both the initial and continuing QA/QC work.
                               JST-42

-------
                                  REFERENCES
1.  Eichelberger,  O.W.,  I.E.  Harris,  and W.L.  Budde,  "Reference  Compound to
    Calibrate Ion  Abundance Measurements in  Gas  Chromatography-Mass
    Spectrometry Systems,"  Anal.  Chem.  47. 995-1000  (1975).

2.  McNair, H.M.,  and E.J.  Bonelli,  "Basic Gas Chromatography,"  p.52,
    Consolidated Printing,  Berkeley,  CA  1969.
                                  IT'43

-------
                  TABLE 1
 OFTPP Key Ions and Ion Abundance Criteria

Mass                     Ion Abundance Criteria
51                       30-60% of mass 198
68                       less than 2% of mass 69
70                       less than 2% of mass 69
127                      40-60* of mass 198
197                      less than 1% of mass 198
198                      base peak, 100% relative abundance
199                      5-9% of mass 198
275                      10-30% of mass 198
365                      greater than 1% of mass 198
441                      less than mass 443
442                      greater than 40% of mass 198
443                      17-23% of mass 442
                  IF-44

-------
                   TABLE 2

   BFB Key Ions and Ion Abundance Criteria



Mass               Ion Abundance Criteria


 50                20-40% of mass 95

 75                50-70% of mass 95

 95                base peak, 100% relative abundance
 96                5-9% of mass 95

173                less than 1% of mass 95
174                70-90% of mass 95
175                5-9% of mass 95
176                70-90% of mass 95
177                5-9% of mass 95
                      - 45

-------
                 TAILING FACTOR= —
                                  AB
Example calculation: Peak Height = DE = 100 mm

                  10% Peak Heights BD =10 mm

                  Peak Width at 10% Peak Height s AC = 23 mm

                         AB = 11 mm
                                         12
                  Therefore: Tailing Factor = — =1.1
                                         11

          Figure 1 .  Tailing factor calculation
                     46

-------
                                   Figure 2
                        Summary of Initial Validation
                       Analyses for Extractable Samples
1.  Composite the Replicate Samples.

2.  Withdraw a 1-liter aliquot.  Store composite at 4°C.
    Separate into three groups of 4 aliquots each.

3.  Determine X the background concentration of each
    Priority Pollutant.

4.  Withdraw twelve 1-liter aliquots from the composite.
5.  a) Dose 2 of aliquots with surrogate standards only at
       100 ug/1.  Label as Al and A2.                         2
    b) Dose 2 aliquots with surrogate standard at 100 ug/1
       and priority pollutants to give a concentration
       level of 2X.  Label as Dl and 02.                      2
    c) Analyze Al, A2, Dl, and D2 using Method 625.

6.  Repeat 5a, b, and c.  Label surrogate standards only
    as Bl and 82.  Use 10X level for priority pollutants.
    Label as El and E2.                                       4

7.  Repeat 5a, b, and c.  Label surrogate standards only
    as Cl and C2.  Use 100X level for priority pollutants.
    Label as Fl and F2.                                       4
                                                           Liters    GS/MS
                                                            Used      Runs
                                                       TOTAL 13        13
                                   JT-47

-------
                                   Figure 3
                        Summary of Initial Validation
                        Analyses for Purgeable Samples
                                                             5-ml
                                                           Syringes   GC/MS
                                                             Used      Runs
1.  Composite the grab sample.
2.  Fill 13 clean 40-ml vials or 3 clean 120-ml vials +
    4 clean 40-ml vials with composite, cap, and store at 4°C.

3.  a) Fill a 5-ml syringe from one 40-ml vial.                1
    b) Analyze and determine X the background concentration
       of each priority pollutant.

4.  a) Fill two 5-ml syringes from one 40-ml vial              2
    b) Dose with surrogate standards at 100 ug/1.
       Label as Al and A2.
    c) Analyze Al and A2 using Method 624.
    d) Fill a 100 ml volumetric to mark using one 120-ml
       or three 40-ml vials.
    e) Dose with 20 ul of priority pollutants to give a
       concentration of 2X.
    f) Fill two 5-ml syringes from the 100-ml volumetric.      2
    g) Dose each syringe with surrogate standards at a
       concentration of 100 ug/1.  Label as Dl and D2.
    h) Analyze Dl and D2 using Method 624.

5.  Repeat 4.  Label surrogates only as 81 and 82.  Use  10X
    level for priority pollutant.  Label priority pollutants
    plus surrogate standards as El and E2.                     4

6.  Repeat 4.  Label surrogates only as Cl and C2.  Use  100X
    level for priority pollutants.  Label priority pollutants
    as Fl and F2.                                              4
                                                      TOTAL     13          13
                                  35T-48

-------
                                   Figure 4
                     Summary of Ongoing Quality Assurance
                            for First Day Samples

                                                              Liters   GC/MS
Extractables                                                   Used      Runs

1.  Composite the Sample.

2.  a) Withdraw three 1-liter aliquots.                          3
    b) Dose two aliquots with surrogate standards
       only at 100 ug/1.
    c) Dose one aliquot ith surrogate standards and
       the priority pollutants, of interest to give a
       concentration of 2X, Section 131.5.
    d) Analyze using Method 625.
                                                 TOTAL          ~3~
                                                                5 ml
                                                              syringes  GC/MS
Purgeables                                                      Used     Runs

1.  Composite the Sample.

2.  a) Fill 4 clean 4-ml vials or 1 clean 120-ml vial
       + 1 clean 40-ml vial with composite.  Store at 4°C.

3.  a) Fill four 5-ml syringes from one 40-ml vial.              4
    b) Dose each with surrogate standards at 100 ug/1.
    c) Fill a 100 ml volumetric to mark using the 120-ml vial
       or three 40-ml vials.
    d) Dose with 20 ul of priority pollutants to give a
       concentration of 2X, Section 132.5.
    e) Fill two 5-ml syringes from the 100 ml volumetric.        2
    f) Analyze two 5-ml syringes containing surrogate
       standards only and one 5-ml syringes containing surrogate          3
       standards and priority pollutants using method 624.

                                                         TOTAL  ~5~      ~T~
                                      49

-------
                                   Figure 5
                          Summary of Ongoing Quality
                     Assurance for Subsequent Day Samples
Extractables
1.

2.
    Composite the sample.
                                                            Liters
                                                             Used
                                                               1
                                                                  GC/MS
                                                                   Runs
a) Withdraw a 1-liter aliquot.
b) Dose with surrogate standards only at 100 ug/1.
c) Analyze using Method 625.                             	
                                         Subtotal        ~~T
                                         TOTAL (x29 days) 29
                                                              5-ml
                                                            Syringes
                                                              Used
Purgeables

1.  Composite the sample.

2.  a) Fill two 5-ml syringes with composite.
    b) Dose with surrogate standards at 100 ug/1.
    c) Analyze one 5-ml sample.
                                             Subtotal
                                             TOTAL (x29 days) 58
                                                                  GC/MS
                                                                   Runs
*Assuming that field blank shows no priority pollutants.
 If field blank were to be analyzed by GC/MS each subsequent
 day, the total would be 58.
                                         50

-------
                                 Appendix A

                          Storage of Final Extracts

Each extract is to be washed out of its container  into a  10-ml  glass  ampul
and brought to 5 ml ^ 1 ml.  Methylene chloride is the solvent  for the
base-neutrals and acids, hexane for pesticides.  The cooled  ampuls (-20°C)
are to be sealed in a rounded-off, fire polished manner,  i.e.,  no thin  sharp
peaks of glass that are easily broken on handling  and shipping.  After
sealing the ampuls, put an indelible mark at the solvent  level.  Securely
attach a label or tag that gives:

    Type of fraction (base-neutral, etc.)
    Industrial category
    Name (of plant, city and state)
    Specific source or stage of treatment
    Date sampled
    Date sealed
    Name of contractor and analytical laboratory

Wrap the ampuls in packing material to prevent breakage and  mail or ship
them postpaid at ambient temperature.  When the samples are  safely in
ampuls, the remainder of the composite sample may be discarded.
                                        51

-------
                                  Appendix B
                    Archiving GC/MS data on Magnetic Tape
When the analyses are completed, transfer GC/MS raw data to magnetic tape as
described below.  All raw GC/MS data is to be retained on 9-track magnetic
tape.  When a tape is filled, deliver the tape to the Environmental
Protection Agency for storage and later evaluation.  The tape format is:
    Type - 9-track, 800 BPI, 2400 foot reels
    Record Length - 80
    Block size - multiple of 80 (specify)
    Code - ASCII or EBCDIC
The data format must be approved by EPA.  At the present time there are
formats used by commercially available equipment that have been approved by
EPA.  Before submitting tapes to EPA, a sample data tape with appropriate
documentation should be submitted for approval of the format and content.

Each GC/MS run on tape should contain the following header information at
minimum.
    1.  Identifying run number
    2.  Data and time of sample run
    3.  Run description.  Include GC column used, temp, program
        (e.g. 70-270 @8), concentration factor of extract, run type
        (e.g. Base/Neutral, Acid, VOA, Direct Aqueous injection).
    4.  Scan time in seconds.
    5.  Mass range scanned.
                                   "TV"- 52

-------
Each scan within each run should contain at minimum:
    1.  Identifying run number.
    2.  Scan number
    3.  Date
    4.  Time of scan (time from start of run).
    5.  Total intensity of scan (not normalized).
    6.   If intensities are normalized, the mass and raw intensity of
         the base peak for the scan.
    7.   Mass/intensity pairs for the scan.  Intensities may be
         normalized or unnormalized.  Normalized values must have a
         dynamic range of at least 1000; unnormalized values must
         reflect the actual absolute intensity value.  Masses may be
         nominal interfer values.
Each magnetic tape reel shall be accompanied by Table of MS response ratios
and a descriptor sheet that includes the following information for each
run.  See Figure GC-MS Transmittal Form, this Appendix.
    1.  EPA sample number
    2.  Laboratory sample number
    3.  Fraction type (base/Neutral, Acid, VOA, Blank, Standard Runs, etc.)
    4.  Industrial Category
    5.  Specific source or stage of treatment
    6.  Date sampled
    7.  Sampler (Contractor or EPA Region)
    8.  Data run
    9.  File number on tape
    10. File name on tape
    11. Comments - to include any deviation from method 624 or 625.
                                     - 53

-------
GC/MS DATA TRANSMITTAL FORM
          (continues on pan,* 2)
1
2
3
4
5
6
7
8
9
10
n
12
EPA Numb*












Lab Numbef












Type Fraction












Inriustnat
Caieqoiy












Specific Sou ice of
Siaqe Ot Tieatment












Date
Sampled












SMnptet-
Contracior 01
EPA Laboratoiv












Date
Run












GC/MS DATA TRANSMITTAL FORM
              i.ine 7
1
2
3
4
5
6
7
8
9
10
11
12
EPA NiMnbtt












Mag Tape
File Numbn












U.iq Tape
Fili> Nnmp












Cuninn*f>t^












    54

-------
                                  APPENDIX C

               Numerical Example of Verification  Phase  Results
The following is an example of the calculations  and  results  of a
verification study.
        Surrogate Standard
            (ug/1)
Sample     Added
 Al
 A2
 81
 B2
 Cl
 C2
 Dl
 D2
 El
 E2
 Fl
 F2
95
95
95
95
95
95
95
95
95
95
95
95
Found

 93
 97
 96
 98
 90
 94
 99
 95
 89
 91
 94
 96
                     Range
                     (Eg. 3)
4

2

4

4

2

2
         Mean
         (Eg. 4)
95

97

92

97

90

95
% Recovery
   (Eg. 10)

    98
   102
   101
   103
    95
    99
   104
   100
    94
    96
    99
   101
    Critical Difference (Eq. 5}  Rc =
    Mean % Recovery (Eq. 8)      P  =
    Standard Dev. of P (Eq. 9)   Sp =
         Acceptable Range of Recovery
                          9.8 at 95 ug/1.
                          99%
                          3.1        3SD »
                           90 to 109X
                                 9.4
During the same verification study, the following  data were
obtained for one of the priority pollutants studied.

               Priority Pollutant                  Percent     Mean
        Back-                    Range    Mean     Recovery  Recovery  Std.Dev.
Sample  ground Added  Found     (Eq. 3)   (Eq. 4)   (Eq. 7)    (Eq. 8)   (Eq.  9)
  X
 Al
 A2
 Bl
 82
 Cl
 C2
 Dl
 D2
 El
 E2
 Fl
 F2
X
X
X
X
X
X
X
12
12
12
12
12
12
0
0
0
0
0
0
0
12
12
no
no
1200
1200
           12
            9
           n
           n
           14
           13
           14
           17
           18
          102
          107
         1160
         1140
             2

             3

             1

             1

             5

            20
        10

        12.5

        13.5

        17.5

       106

      1150
       42
       50
       82
       86
       96
       94
      46

      84

      95
5.9

3.3

1.2
                                       55

-------
    Critical Difference    (Eq. 5)  Rc =  6.7 at 12 ug/1
    for priority pollutants         Rc s  3.3 at 17.5 ug/1
                                    Rc =» 16.4 at 106 ug/1
                                    Rc = 65 at 1150 ug/1

    Mean Value of X  (Eq. 6)          X = 12.0 ug/1 (background concentration)

Is there a significant difference in the recoveries between the 100X and  10X
levels?  Apply a two tailed student's t-test with a confidence level of 95%.
Sd -  MoO- SlO2  - 7(1.2)2+ (3.3)2  = 3.51
     tf              v                 2 degrees of
                                                 degrees of freedom

                  - *10  -  95-84 = 11

         texp  =  d/Sd = 11/3.51 = 3'. 70

Since 3.7 is less than 4.3 (t-value, 0.95, 2 degrees of freedom) there  is no
significant difference between the 100X and 10X levels.  Apply equations 8
and 9 to the four recoveries for these two levels.  The mean recovery (eq.8)
is 89% with a standard deviaion (eq. 9) of 6.5 (3 degrees of freedom).  Test
the 2X level against this mean recovery and standard deviation.

         Sd  = ,/(6.5)2 + (5.9)2  = 8.78 (4 degrees of freedom)

         d  =  89-46 * 43

               «  43/8.78  =  4.90
Since 4.90 is greater than 2.78 (t value, 0.95, 4 degrees of freedom) there
is a significant difference between the 2X and the 100X, 10X levels.

                                       2X        100X, IPX
         Mean % Recovery (Eq. 8)       "41T            §9
         Standard deviation (Eq. 9)     5.9           6.5
         3 Sn                        18            20
         Acceptable Range (%)        28 to 64      69 to 109


During day one of the 30-day study the following data were obtained.

         Surrogate Std.                    Priority Pollutant
           (ug/1)                              (ug/1)

                         Recovery                                % Recovery
Sample   Added   Found   (Eq. 13)          Added       Found      (Eq.  14)

  a      100      93       93                0           13
  b      100      90       90                0           15
  c      100     105      105               12           20          50%
                                  T3T 56

-------
For the Surrogate Standards, on day one, the range between  a  and b  (Eq.  11)  is
3, and the mean concentration (Eq. 12) is 91.5 ug/1.  Since the critical
difference is 9.8 at 95 ug/1, this range is acceptable.  The  recoveries-of  the
Surrogate Standards in a, b, and c (Eq. 3) are all in the acceptable range  of
90 to 10955.  Therefore, the accuracy is acceptable.

For the Priority Pollutant, on day one, the range between a and b is
acceptable, (less than 6.7) and the recovery is acceptable  (acceptable range
of 28 to 64*).

The following is an example of the results for day 2 to 30 of the 30-day  study:
              Priority
            Pollutant (ug/1)
     Day      Found

      2         14
      3         15
      4         20
      5         11
      6         17
      7         18
      8         10
      9         14
     10         13
     11         12

     12         20

     30         15
Surrogate Standard (ug/1)
Added   Found   % Recovery
 100
 100
  00
 100
 100
 100
 100
 100
 100
 100

 100

 100
 90
 99
100
107
100
100
 75
 92
 93
 94

 95

 95
 90
 99
100
107
100
100
 75
 92
 93
 94

 95
-Update Recovery for SS*

-Not Acceptable
   Disregard
   Results
•Update Recovery for
   SS**
 95 -Give final statistic on
       Recovery of SS.
*  Includes 20 results, 12 validation, 3 first day and days 2 through 6
     Mean =99  Std. Dev. = 4.6   Acceptable range 85-113%

** Includes 25 results; day 8 result not included.
     Mean = 98  Std. Dev. = 4.6  Acceptable range 84 -112%
                                       57

-------
                                  Appendix 0
                        Possible Sources of Standards

Aldrich Chemical Co., Milwaukee, Wise.;
Analabs, Inc., North Haven, Conn.;
J. T. Baker Chemical Co., Phillipsburgh, N.J.;
Chemical Procurement Laboratories, College Point, N.Y.;
Columbia Organics Catalog A-7," Columbia, i.C.
Eastman Kodak Co., Rochester, N.Y.;
K&K Rare & Fine Chemicals, Plainview, N.Y.;
Nanogens International, P.O. Box 487, Freedom, CA  95019
Pfaltz & Bauer Chemical Co., Stamford, Conn.;
RFR Corp., Hope, R.I.; "Chemical Standards for Air-Water-Industry-Foods"
(1975).
"Analytical Reference Standards and Supplemental Data for Pesticides and
Other Selected Organic Compounds," EPA-660/9-76-012 (May 1976), Health
Effects Research Laboratory, Environmental Toxicology Division, Research
Triangle Park, NC.

R K Chemical Co., 2135 Howard St., Hartville, OH  44632
Tridom Chemical Inc., Hauttauge, N.Y.
NOTE:  These sources are not to be interpreted as being endorsed by the EPA.
                                    J7-58

-------
                                   GLOSSARY

EICP - Extracted  Ion Current Profile.  EICP is a reduction of mass
     spectr^/aetrfc data acquired by continuous, repetitive measurement of
    spectra that plots the change in relative abundance of one or more ions
    as a function of time.

Equivalent -  A piece of apparatus, gc liquid phase, etc. is
    equivalent if its performance gives results identical to or better than
    that specified.  Contact Equivalency Staff, EMSL, Cincinnati for guide-
    lines in equivalency for Alternate Test Procedures.
FID -  Flame lonization Detector

Field Blank -  Organic-free water that has been transported to the
    sampling site and back to the analytical laboratory in a container and
    bottle identical to the type used to collect the samples.  Field blanks
    and samples must be shipped in separate containers.  When received in
    the lab, the Field Blank is dosed, extracted and concentrated as if it
    were an actual sample.

Limit of Detection -  10 ug/1 unless otherwise specified.  For the
    GC/MS Methods, 624 and 625, this is an operational limit that must be
    met.  The entire analytical system must give recognizable mass spectra
    and calibration points from standards representing this concentration.
    For Method 624, the condition (at 10 ug/1) must be met when sparging 50
    ng of a component from 5.0 ml of water.  For Method 625, the conditions
    (at 10 ug/1) must be met when injecting standards at a concentration of
    10 ug per ml.

Method Blank -  For Method 625, a 1-liter volume of organic-free water
    that is extracted and concentrated to the final volume of 1 ml.  For
    Method 624, a 5.0 ml aliquot of organic-free water.

Organic-Free Water -  Water that gives no interferences when analyzed
    as a Method Blank.   See Method 624, Section 5.4, for the preparation of
    organic-free water.

Sample Blank -  Identical to Field Blank.

Significant Difference -  Check for difference in magnitude by using a
    two-tailed Student's t-test at a 95% Confidence Level.

Significant Drift -  A 25% or larger change from an averaged
    response.  Must be based on a constantly updated average.
                                   IT- 59

-------
SIM -  Selected Ion Monitoring - Measuring  the  mass  spectrometr*) c
    response to one or several iojis  in  real  time.
Surrogate Standard -  Compounds not expected  to occur in the
    environmental sample that  are  added  to  the sample to monitor for
    unusual matrix effects, gross  sample processing errors,  etc.  The
    compounds chosen should be chemically similar to the pollutants
    being measured.  Surrogate Standards are  not used as internal
    standards.

Transfer Pipet -  Class A, to  deliver, pipet.
              RIMmMQfTICt-Bn

-------