United States      Office of Air Quality       EPA-450/3-83-020
           Environmental Projection  Planning and Standards      December 1983
           A9encY        Research Triangle Park NC 2771 i
SEPA      Revisions to
           Methods 4 and 5,
           Appendix A
           of 40 CFR Part 60 —
           Summary of Comments
           and Responses

-------
                                        EPA-450/3-83-020
      Revisions to Methods 4 and 5,
    Appendix A of 40 CFR Part 60 —
Summary of  Comments and Responses
             Emission Standards and Engineering Division
            U.S ENVIRONMENTAL PROTECTION AGENCY
               Office of Air, Noise, and Radiation
             Office of Air Quality Planning and Standards
            Research Triangle Park, North Carolina 27711

                    December 1983

-------
This report has been reviewed by the Emission Standards and Engineering Division of the Office of Air
Quality Planning and Standards, EPA, and approved for publication. Mention  of trade names or
commercial products is not intended to constitute endorsement or recommendation for use. Copies of
this report are available through the Library Services Office (MD-35), U S Environmental Protection
Agency, Research Triangle Park, North Carolina 2771 1; or, for a fee, from the National Technical
Information Services, 5285 Port Royal Road, Springfield, Virginia 22161
                            Publication No EPA-450/3-83-020

-------
                            TABLE OF CONTENTS





                                                                 Page
Chapter 1.  Introduction
Chapter 2.  Summary of Changes Since Proposal ..........   2
Chapter 3.   Summary of Comments  and  Responses ..........   3
     Table 1.   List of Commenters   ...............  35

-------
                                 Chapter 1

                                INTRODUCTION

      On September 7,  1982,  the U.S.  Environmental  Protection Agency
 (.EPA) published in the Federal  Register (47 FR  39206)  "Revisions  to
 Methods 4 and 5."   These revisions  were proposed  under  the authority of
 Sections 111, 114,  and 301(a) of the  Clean  Air  Act  as amended.
      Public comments were solicited at  the  time of  proposal.  To  provide
 interested  persons  the opportunity  for  oral  presentation of data, views,
 or arguments  concerning  the proposed  revisions, a public hearing was to
 be held,  if requested, after November 8, 1982.  The hearing was  not held
                    »
 because  no  one  requested  to speak.  The public comment period was from
 September 7,  1982, to November 8, 1982.
     Twenty-two .comment letters on the proposed revisions were received
 from industry, State air pollution control  agencies, trade associations,
and testing contractors.  The comments that were submitted, along  with
EPA's responses, are summarized  in this  document.  The  summary of
comments and responses serves  as a basis for the revisions  that have
been made to the proposed revisions  between  proposal and  promulgation.

-------
                                Chapter 2

                    SUMMARY OF CHANGES SINCE PROPOSAL.

     1.  Method 2, Section 4.2.   The quality assurance procedure has
been changed to a quality control  (optional)  procedure and  moved to
Section 2.2.7.
     2.  Method 5, Section 4.4.   The quality assurance procedure has
been changed to a quality control  (optional)  procedure.
     3.  Method 5, Section 4.4.1.   The procedure  has  been changed
from an orifice factor determination to a check of AH@ and  meter
calibration.

-------
                                  Chapter 3
                      SUMMARY OF COMMENTS AND RESPONSES
  Commenter IV-D-l
       1-1  Comment:   The first subparagraph  entitled  "Orifice Factor
  Determination"  should be labeled  "4.4.1"  rather than 4.4.
            Response:   This  is  corrected  in the promulgation version.
       1-2  Comment:
            a.  The units  "mm H20" must appear before "(in H20)" in the
  definition of aHj.
            b.  The empirical constants (a and b)  of paragraph  4.4.2  can
  account  for unit conversions in both Qj  and  Orifice Factor (OF).
  However, if a standard temperature and pressure  conversion factor is
 desired, paragraph 6.3 defines K which could be  used.
           c.   Except  in paragraph  6.1, Tm  is  not defined.   But using
 6.1's definition causes the equation  to  be invalid.  The  "+273" should
 be removed from  the denominator.
           Response:   These  are  clarified in the promulgated version.
      1-3   Comment:  Equation numbers  for OF and the'regression equation
 of paragraph 4.4.2 should be added.
          Response:  Although these equations are no longer in the
 revisions, the new equations are numbered.
     1-4  Comment:  In the Qt  equation, is  the sign minus?
          Response:   The sign should be plus.
     1-5  Comment:  Can undried air be properly used to establish the
regression equation?   The field check  procedure of  paragraph 4.4.3 may

-------
imply that the itnpinger train is connected,  thereby,  presenting  dried
air to the dry gas meter.   Does this  make  a  difference?
          Response:  This  should not  make  any  difference  as  the
relationship between the orifice reading and the  volume measurement
will remain the same.
Commenter IV-D-2
     2-1  Comment:  The sign in the numerator  of  the  equation  in
Section 4.4.1 should be a  plus.
          Response:  We agree.
     2-2  Comment:, The molecular weight of  air should appear  in  the
OF equation.
          Response:  Air in the gas was used for  both the calibration
and the field check and was considered a constant.  Therefore, it was
not included as a variable in the equation.
     2-3  Comment:  "4.4  Orifice Factor Determination" should read:
"4.4.1  Orifice Factor Determination."
          Response:  We agree.
     2-4  Comment:  It is not clear why the  Environmental Protection
Agency  (EPA) has suggested fitting the data  to a  straight line.   After
calibrating  a meter box against a wet test meter, a test  firm  has a table
of  gas  volume data and AH data for the meter box.  This tabulated
data can  be  taken  into the field and compared  with the  field test
calibration  check.  The test  firm could operate the meter box  at a
specific  AH  (i.e., a AH for which there is a corresponding gas

-------
  volume in the tabulated data), and measure the resulting gas  volume.
  This gas volume can be compared to the gas  volume  measured at  the  same
  AH during the calibration in the lab  to see if  it  meets  the  1.5
  percent criteria.   There is  no  need to  actually  plot the data or to fit
  the data  to  a straight line.   It is also  not  clear why 1.5 percent and
  not 2.0 percent  was  selected  as  the criterion, since Method 5 specifies
  a  +2.0  percent accuracy  for  the  gas meter.
           Response:  We  agree, although, because of differences in
  ambient conditions,  even ±2 percent may be too strict.   It  is possible
  to  perform the calibration at one point and  repeat  the  check  at the same
  point with corrections for condition differences.   This is  the  approach
  in the final  version.
 Commenter IV-D-3
      3-1  Comment:   The commenter restated the intent of  the  proposal
 and noted  that the  changes  should be beneficial.
           Response:   No response  necessary.
 Commenter  IV-0-4
      4-1   Comment:  The proposed  revisions to Method 4 and Method 5,
 contained  in Appendix A of 40 CFR Part 60, is in  conflict with allowable
 pretest  calibration limits and adds little if any quality assurance
 (QA)  to  the two methods.  The revisions,  as written,  would remove  the
 calibrated wet test meter as the primary  standard  and  replace  it with
 the orifice.   The use of a calculated OF  for  a  single point
calibration verification in  the field (±1.5 percent  flow rate  error)

-------
requires a stricter equipment error tolerance than that required for
pretest calibration of the sampling equipment.   The single point
calibration check must then be subjected to a best-fit curve,  obtained
by linear regression,  which contains an inherent error due to  its  nature
of being "best-fit."
     The commenter continues with an example using actual  calibration
data and assumed data  errors to show the effect of errors  on the OF.
     The QA procedures,  outlined in Volume III,  seem adequate  with
respect to the existing  equipment available for stack  sampling.  As
technological  advances improve the quality of the equipment, the
accuracy of the sampling methods will  increase  as  well.  Therefore,  it
is recommended the proposed field +1.5  percent  flow rate QA check  not be
incorporated into Method 4 and Method 5.
          Response:   The tolerance limits  on the OF are somewhat
restrictive for field  applications.   The final  version  of  the  quality
control (QC)  has greater tolerances of acceptance of +3 percent.
Commenter IV-D-5
     5-1  Comment:   The  recent proposal  (Federal  Register  of
September 7,  1982)  to  require a field check  of  the dry  gas meter via
orifice meter  for Methods 4 and 5 appears  to have little practical value.
     I can see that there is some merit in having  the  proposed
correlation line equation on hand if the person  performing the test
suspects his  equipment to be faulty.   However,  one can  get a fair
indication of  how things are going after the first test run if the

-------
                             )
 moisture collected  is measured  and  then  the percent  isokinetic value is
 calculated.   If  the  initial  moisture  guess approximates what is
 measured and  if  the  percent  I then  falls within the  acceptable range,
 that  is  an  indication that the  dry  gas meter is in order (I assume the
 tester is conscientious  in taking data readings and  adjusting sampling
 rates).   And,  some of the necessary work will  be done.
      This pretest check  will make no  difference in the accuracy of the
 test  results.  Accuracy  depends on  the test equipment calibration
 performed before and after testing  as well as on the care given to
 proper sampling, cleanup, and laboratory procedures by the persons
 involved.
      All  that will come  from such checks is that,  on rare occasions,
 someone will avoid going through the motions  of a  test only to  have to
 scrap the results because of a faulty dry gas  meter.   You certainly
 could not invalidate a test after the fact because of the absence of
 this  field calibration check since it is in no way involved in
 calculating the results.
      I expect the precheck procedure will  seldom be performed (if
written into the method)  unless  a government  inspector demands  it.  In
most  all  cases, such an  inspector will be viewed as tedious and obtuse.
We do not need the ill will.
          Response:   The  commenter mentions some good equipment
specifications and checks that are already  required by  Method 5 and
other regulatory  procedures.   However, these procedures are not on-site

-------
quality control  (QC)  checks.)  The intent  of a QC check is to prevent the
collection of inaccurate sampling data.   The QC check before the test can
save considerable amounts of time if  faulty equipment is discovered
before the test.   We  agree that  such  a  QC check should be optional, but
it is a recommended one.
Commenter IV-D-6
     6-1  Comment:   There appears to  be an improper  relationship in the
published equations.   Also,  the  1.5 percent criteria is considerably
more restrictive than the existing criteria of 5 percent between
pre- and post-test meter calibrations presented in Section 5.3
(42 FR 41781).  A reference regarding the application of this QA
procedure to field data would be helpful.
          Response:  The equation did not account for small differences
in temperature or pressure which can  cause the calculations to  show
greater than 1.5 percent differences  without  the meter being out of
calibration.  See also Responses 2-4  and  4-1.
Commenter IV-D-7
     7-1  Comment:  The proposed revisions would require the source
tester to perform a field check  of the  dry gas volume meter versus the
OF prior  to a series of emission tests.  Methods 4 and 5 currently
require extensive calibration procedures  which must  be conducted before
and  after the metering  system's  use in  the  field.  The calibrations
must differ by no more  than 5 percent for the sampling results  to  be
acceptable.  The existing calibration requirement  is more  than

-------
                             I
 sufficient  to  guarantee  the accuracy of the method.  In the absence of
 strong  supportive data,  we fail to see that the additional time and
 trouble involved with the implementation of this revision would be
 justified by any associated benefits.
          Response:  We  agree.  The QA procedure required for each test
 as  stated in the proposal has been revised and designated a QC
 (optional)  procedure.  See also Response 5-1.
 Commenter IV-D-8
     8-1  Comment:  We applied the proposed QA procedures to actual
 equipment operation and  found a contradiction  between the existing
 QC regulations and the proposed audit check.   The field check  resulted in
 a 2.6 percent deviation  from the predicted flow rate, which  would  require
 recalibration of the metering system under the proposed regulation.   A
 Method  5 test was then performed,  after which  the equipment  was  brought
 back from the field and  the post-test calibration check required in
 Section 5.3 of Method 5 was conducted.   The unit  passed this post-test
 check with a 3 percent deviation from the  Y factor calculated  during  the
 pretest calibration,  well within the  limit  of  5 percent.
     The above discussion shows that  the proposed field check  can  result
 in needless  recalibration of equipment  that passes  the  QC  procedures
currently included in Method 5.  The  figure of 1.5  percent deviation
between actual  and predicted flow  rates  as  a test  of  acceptability

-------
appears to be extremely arbitrary and restrictive.   The QA procedure
proposed in the regulation must be no more restrictive than the QC
function it is designed to audit.  The criteria for acceptance must be
re-examined to provide a realistic, attainable tolerance level  which'
will provide a check of the calibration and not a completely separate,
independent, and arbitrary level  as is now the case.
          Response:  See Responses 5-1 and 7-1.
     8-2  Comment:   An alternative procedure,  using a  critical  orifice,
is allowed if approved by the Administrator.   Such  a procedure would
undoubtedly be quicker and easier than that specified  in the proposed
rules, and as such  would be more  readily  accepted than  that and
implemented by the  regulated community.   But the  process  of gaining
approval for each facility which  wants to use  this  procedure could  be  so
lengthy and involved that few facilities  may take this  route.   Therefore,
we request that EPA develop this  alternative procedure  and  promulgate
guidelines for its  implementation.
          Response:  The critical  orifice approach  to calibration check
of the dry gas meter remains an alternative procedure.   The procedure is
under development within the Agency,  but  is not yet ready for universal
use.
Commenter;   IV-D-9
     9-1  Comment:   The minus sign in the volume  flow rate  equation  in
Section 4.4 should  be a plus sign.
          Response:  See Response 2-1.
                                    10

-------
       9-2   Comment:   The meter factor,  Y,  should  be  the  factor, Yj,
  during  run  j  instead  of the  average  Y  obtained during the calibration
  procedures.
           Response:   We agree.
      9-3  Comment:  This proposal is unnecessary based on Alcoa's use
  of control consoles.  The Environmental Control Laboratory at the Alcoa
  Technical Center (ATC)  is continuously shipping control  consoles to Alcoa
  plant locations all over the country for environmental  sampling.   Upon
  return to the ATC laboratory, the units are calibrated,  and  there are no
 problems with the units meeting the  calibration requirements  of  Y =  l.OQ
 + .02 and AH@ = 1.83 + .25.
           Response:   The criteria mentioned  by  the commenter  are  not
 accuracy  checks,  but are equipment specifications.   Post-test agreement
 with  such  specifications are  good indications of  properly operating
 equipment, but is not  the same as  a field QC check.  See also Response
 5-1.
      9-4  Comment:   If a  ruling is made requiring field checks on
 control consoles, the proposed method does not fill  this  need as  seen
 from  the field checks attached to this letter,  where a new console
 calibrated and checked on the same day,  without moving,  showed a  2.8
 percent - 4.2 percent difference in predicted vs.  actual  readings.  A
more reliable field  check should be explored.
                                    11

-------
          Response:  The Agency reviewed these data and can find no
obvious errors, but note there are two other sets of data showing the
checks are within the criterion.  There are several possible factors
that could have caused such differences:  significant temperature
differences, inconsistent technique, or malfunctions of the metering
system.
     9-5  Comment:  The^1.5 percent between the measured and predicted
flow rates is too restrictive.  For standard calibration,  the meter
coefficient of the control  console is acceptable at 1.00_+ 0.02.   As
stated in Docket No., A-82-06,  Method 5,  Section 7.1.1.5,  the meter
coefficients should be between 0.95 and 1.05 for a  dry gas meter used as
a calibration standard.   The acceptable percent is  not consistent for
these three cases.  From the attached data  and the  calibration standard,
it would seem that the difference  between the measured and the predicted
flow rate should be at least_+5 percent.
          Response:  See Response  9-3.
Commenter IV-D-1Q
     10-1  Comment: The proposed changes are discriminatory and place an
undue burden on sites that  conduct their own emissions testing.   The
current regulations require that the dry gas meter  be calibrated versus
a wet test meter before and after  every test.   The  proposed changes
would require additional  field calibration  testing  after  moving the
equipment several feet to the  test site. This appears to  be redundant
and burdensome to those firms  which have invested capital  and calibrate
versus a wet test meter on  site.
                                    12

-------
            Response:   We agree.   The  proposed procedure was not intended
  to  supersede  other valid  calibration procedures and in the promulgated
  version  is  an optional  one.
  Commenter IV-D-11
      11-1   Comment: The implementation of these revisions is  unwarranted
  and constitutes an unreasonable burden. Further, that benefits realized
  to the environment would be minimal,  if any at  all.
      Essentially,  EPA's proposed revisions would involve  a  detailed  and
 lengthy procedure,  which attempts to  check meter volume to  an  extremely
 precise degree (1  1/2  percent),  while the EPA's  own  study of Method  5
 precision indicated a  standard deviation  of  31.1 percent
 (EPA-650/4-74-013).
           Response: See Response 5-1.   In  addition, the commenter cites
 a result  of  a  much earlier collaborative  study of Method 5 conducted at
 a power plant.   The quoted standard deviation refers to within laboratory
 precision of particulate sampling results.  This is not the same as  the
 precision that can be  achieved in calibrating the dry gas  meter.
 Repeated calibrations with EPA Method 5 equipment have demonstrated  that
 the 1 1/2 percent limit can be met;  however,  the limit suggested  for  the
 field check  has been raised to 3 percent.   The Agency  feels  that this
 degree of precision in  the field is  a  good indication  of the proper
operation  of the volume meter.
     11-2   Comment:  The improvement  in  test  precision would be
negligible,  due to  other large  inherent  sources of error.
                                   13

-------
                             r
          Response:  The procedure is a QC measure - a check of system
integrity.  It will not improve precision of the test data,  except to
indicate those results collected with faulty equipment.
     11-3  Comment:  The additional  testing time involved (with the
unit off economic dispatch) is estimated to be 1 and 1/2  hours.  The
Florida Power and Light Company has  estimated the additional  cost to
our rate payers to run one of our new units in this  manner to be
between $24,000 to $100,000 per unit/day.
          Response: This comment indicates a misunderstanding of the
intent.  See Response 5-1.
     11-4  Comment:  A Florida Department  of Environmental Regulation
post-test calibration check is  already  routinely conducted.
          Response:  See Response 5-1.
Commenter:   IV-D-12
     12-1  Comment:  A pretest check  would mainly  increase the
tester's confidence in the  volume metering system  prior to testing.   If
the metering system has not been abused  or damaged since  its  last  field
use, the post-test calibration  check  of  the volume metering  system
after its last field use (as specified  in  Section  5.3) usually  assures
the tester of  the accuracy  of the metering system.   Any pretest  check
of the volume  metering system should, at most, be  recommended  at  the
tester's discretion similar to  the Method  5  recommendation of  a
pretest sampling  train leak-check.
          Response:   We agree.
                                    14

-------
     12-2  Comment:   Section 4.4, Orifice Factor Determination,
should be renumbered to Section 4.4.1.   In accord with  the  conventions
already used in Method 5, the equation  for calculating  volume  flow  rate
should be numbered and corrected.  In a similar manner,  the orifice
factor, OF, equation should be numbered,  and the term OF in the  equation
should be rewritten as OF-j.
          Response:   See Responses 2-3, 2-1, and 9-2.
     12-3  Comment:   Section 4.4.2 should have  the equations used to
determine "a best-fit linear regression equation ..." specified.  All
mathematical calculations required by Method 5  are presented within
Method 5.  Specification of the necessary mathematical  equations
eliminates a tester's confusion as to the appropriate calculations  and
eliminates potential  conflicts between  the tester and the observing
agency over the methods used.
          Response:   The promulgated version of the procedure  does
not require the use of this equation.
     12-4  Comment:   In Section 4.4.3,  the use  of the word  "run" is
confusing and should be replaced with "check run."  This  would eliminate
the possible interpretation that this check should be performed  following
each emission test run.
          Response:   We agree.
     12-5  Comment:   Further, Section 4.4.3 states,  "The  meter
coefficient is determined to be acceptable if the measured  flow rate
                                    15

-------
agrees with the predicted flow to within +1.5 percent of the measured
value.  If this criterion is not met,  the meter sha'P  be recalibrated
prior to use as described in Section 5.3."  We feel  ^hese sentences
should be replaced with:  "The meter coefficient is  determined  to  be
acceptable if the measured flow rate agrees  with the predicted  flow to
within +_5.0 percent of the measured value.   If this  criterion is not
met, at the discretion of the tester,  the meter sha'M  be replaced  for
the test run series or recalibrated after field use  as  described in
Section 5.3."  We feel our changes are valid for the following  reasons:
    'a.   A criterion of +1.5 percent agreement between  the measured and
predicted flow rates is more stringent than  the +5.0  percent criterion
specified in Section 5.3 for the post-test meter calibration check.
There is no rationale for making the pretest check of the OF more
stringent than the post-test requirements.
     b.   There is no published data base to  show that the specified
criterion of a +1.5 percent agreement  between  the measured and  predicted
flow rates can be met.
     c.   The proposed revision is intended to  check  the  accuracy of the
volume metering system calibration coefficient by determining the
accuracy of the OF at the metering system's  AH@.  Since  the orifice is
not an integral  component of and is located  after the dry gas meter, the
accuracy of the dry gas meter is not necessarily  dependent on the
accuracy of the orifice; i.e.,  the orifice's  accuracy may change without
                                    16

-------
  a corresponding change in the accuracy of the meter calibration
  coefficients.   However,  a significant  change  in  the meter calibration
  coefficient without a  significant  change  in the  OF  would require an
  independent, compensatory,  and highly  Improbable change in the orifice.
  Thus,  if  the OF is  accurate,  then  the meter calibration coefficient is
  most probably correct, although, the converse is not necessarily true.
      d,   A  difference between the measured and predicted flow rate
  merely indicates that a problem may exist with the volume metering
  system.  Since  the metering system's AH@ is only used to establish
  the isokinetic  sampling rate relationship  which has  a criterion of J;10
 percent,  we feel the rejection of the volume metering systems which fail
 to meet the orifice  check  criterion prior  to recalibration is much too
 stringent  and also  unnecessary.
     Rather, if  the  pre-test OF check does  not meet  the  specified
 criterion,  then  the  tester should have the  option of  using the metering
 system  for the emission test runs and recalibrating  the metering system
 after the  emission test runs as outlined in Section 5.3 or replacing  the
 metering system  before beginning the emission test runs.
          Response:   We agree.   See also Responses 5-1 and 7-1.
     12-6  Comment:   For over 5 years, we have  checked the accuracy
of our metering  systems'  AH
-------
equations to establish the relationship between Q and OF are eliminated.
Further, the use of a nomograph in the field simplifies  and  expedites
pretest orifice checks by eliminating the mathematical calculations
required in the proposed revision.  To use the enclosed  nomograph, only
1  cubic foot is pulled through the metering system;  however,  the
nomograph can be easily modified for greater sample  volumes.  We  suggest
that the EPA propose the use of a nomograph as an alternative to  their
proposed orifice check and that the nomograph  to  be  used  be  published
along with instructions for its use.
          Response:   We agree.  This  is essentially  the  approach  we have
taken in the promulgation version of  the QC check, although,  we have not
included the use of the nomograph.  An equation  is included  instead.
Use of a nomograph is, of course, acceptable.
     12-7  Comment:   Section 5.3 of reference  Method  5 states, relative
to post-test meter coefficient checks, "Alternative  procedures, e.g.,
using the orifice meter coefficients,  may be used, subject to the
approval of the Administrator."   By using the  linear  relationship
between flow rates and orifice factors,  as  specified  in  the  proposed
revision, and the actual  field data from each  emission test  run,  an
alternative procedure for volume metering calibration checks  can  be
proposed.  The alternative procedure  can be used  in lieu  of post-test
neter coefficient checks.
                                    18

-------
      Using the OF value, the predicted flow rate can be calculated from
 the linear regression equation for the volume metering system; the
 actual flow rate is calculated with the flow rate, Q, equation in Section
 4.4.1 of the proposed revision.  If the measured and predicted flow
 rates agree within ^5 percent,  then the volume metering system is
 acceptable.   Should the flow rates differ by more than +5  percent,  then
 either a post-test meter coefficient check or a full  recalibration  of
 the volume metering  system  must be performed.
      We feel  this alternative to  post-test meter coefficient  checks has
 several  important advantages.   Actual  field data collected  during
 emission  test  runs would be used  to  check  the accuracy  of the  volume
 metering  system.  The  time  and  expense associated with  post-test meter
 coefficient checks would be markedly reduced.  All emission test
 reports,  where emissions were determined with Method(s) 4 and 5, should
 contain all the data necessary  for the report reviewer to perform the
 orifice factor checks  for each  emission test run.
          Response:  This post-test check of orifice coefficient does
 not check the metering system for leakage which is a major  cause of
metering  inaccuracy.  The post-test calibration in Method 5  is
sufficient and no new post-test  checks  will be added.
Commenter IV-D-13
     13-1  Comment:   This proposed revision would add  time  requirements
to already very time-consuming source test procedures,  with  minimal QA
benefits.   Similar QA benefits could  be achieved by calibrating in the
                                    19

-------
laboratory prior to field testing, rather than in the field,  with any
calibration problems identified and resolved before commencing expensive
field sampling activites.  We see no justification for performing the
calibration check in the field.
          Response:  See Response 5-1.
Commenter IV-D-14
     14-1  Comment:  We feel  that this  proposed QA field check will
provide valuable data to both testers and observers.   We feel  that the
use of calibrated orifices would serve  as a more independent  check of
data for test observers and should be used as  the primary technique.
          Response:  See Response 8-2.
Commenter IV-D-15
     15-1  Comment:  This division agrees that this is an excellent QA
tool for organizations who conduct a number of source tests without
recalibration of the dry gas  meter.   However,  this division conducts
only a limited number of source tests,  and it  has been our internal
calibration policy since 1974 that the  dry gas meter be recalibrated
with a wet test meter after one or,  at  the most,  two source tests.  A
number of industrial firms within the State that conduct source testing
follow a similar procedure.   Therefore,  it is  felt that the frequent
recalibration with the wet test meter makes the QA procedures  outlined
in this section unnecessary,  and it  is  requested that these be made
optional for situations where there  is  a frequent recalibration of the
dry gas meter.
                                    20

-------
           Response:  See Response 10-1.
 Coironenter IV-D-16
      16-1  Comment:  Any procedure which has the potential of cancelling
 a scheduled emission test hundreds of miles from the testing crew's home
 office becomes an important consideration to all persons involved in
 that activity.  The proposed QA check is certainly meritorious  in its
 objective of assuring a continuing and known relationship between the
 dry gas meter and the metering orifice installed downstream  thereof.
           Response:   Ho response is  necessary.
      16-2  Comment:   After  reviewing  various  aspects  of  this  proposed
 check procedure,  I have concluded  that from my point  of  view  a more
 simple  procedure  would  be recommended  or  lacking  the  adoption of  the
 suggested alternate  procedure,  then the percentage  variation  allowed by
 the  EPA proposed  method should  be  broadened to approximately +3 percent
 deviation.  The rationale and considerations leading  to  these
 recommendations are  discussed in the following text.
      It  is my  position  that the proposed EPA QA check method is
 cumbersome.  The development of a least squares derived slope intercept
 equation  relating standard cubic feet per minute (SCFM) to the OF is
 really not a problem with the advent of modern computers  and  programmable
 calculators on the market.   The cumbersome aspect derives in  the  field
utilization of this equation wherein  the operator must first  determine
his actual cubic feet per minute (ACFM)  for a 10-minute period, a
procedure which is not suggested to be  cumbersome,  and then convert  this
                                    21

-------
ACFM to an appropriate SCFM figure.  In making this calculation,  he
should refer to some written reference for the appropriate  equation and
the appropriate gas meter coefficient Y in order'to get his measured
SCFM.
     In addition to this calculation, he must calculate the on-site OF
for entry into the slope intercept equation from which  he derives  a
predicted SCFM.  These two SCFM values are then compared to determine
compliance with the proposed j;l.5 percent limitation.   This activity is
certainly not beyond the ability  of a qualified emission testing crew;
however,  it does add to the time  and complexity of  a total  emission
test.
          Response:  See Response 12-6.
     16-3  Comment:  In many areas of the country and specifically  in
the State of Florida, a more simple method has been  adopted for
accomplishing somewhat the same purpose.   In  this simpified method  the
operator determines an on-site ACFM having operated  the meter  box  at the
Meter Box AH@ for a period of 10  minutes.   This measured ACFM  is
then compared to 0.75 CFM which is the value  that an orifice displacement
equal to  AH@ should produce.
     If the measured ACFM is within _+5 percent of 0.75  CFM,  then the
meter box is considered to be in  satisfactory condition, and the test may
proceed or be considered accurate in the case of a  post-test QA check.
It is suggested to you that this  procedure accomplishes the same basic
purpose as the one proposed in the Federal  Register.  The_+5 percent
                                    22

-------
 limitation might be questioned as being somewhat lenient;  however,  it
 does seem to produce satisfactory results.   I  would also  suggest  that  in
 the overall scheme of things as relates to  Method 5 sampling  a _+5 percent
 limitation would seem to be in keeping with the  accuracy  and  capability
 of the entire system.  The computations involved in this  method are
 certainly much easier in the field  and do not  require  reference to any
 written material  to determine compliance or noncompliance.
           Response:   This  approach  is  the one  included in the
 promulgated version of the QC  check, although, the  recommended limit
 will  be +3 percent  rather  than 5  percent.   See also Response 12-6.
      16-4   Comment:   It  is  suggested that there are problems in
 determining  a  slope  intercept  equation  by least squares fit using  a
 small  number of data  points  (five or six in this case)  along with  the
J;1.5  percent limitation  at the end of this procedure.  The sensitivity
 on the  part of the  resulting slope intercept equation to small changes  in
 one or  more coordinates  of the limited data  points suggests  to me  that
 the +1.5 percent limitation is perhaps too restrictive.  I  can foresee
 cases where a testing organization could have what would  otherwise be a
 perfectly good set of calibration data; however,  they would  never  make
 tnej;1.5 percent check by virtue of  the fact that one outlier  among their
 five or six data points is unduly influencing the least squares
determined slope intercept equation.
          Response:   We agree and have  recommended ^3 percent  as a
1 i mi t.
                                    23

-------
      16-5  Comment:   In  connection  with  this  set of comments,  I have
 surveyed a number of emission  testing  agencies  in our local area with
 respect to their calibration practices and ability to meet the proposed
 EPA QA check.   It has  been my  observation that most, of these agencies
 (including mine)  have  been calibrating meter boxes using a wet test
 meter  having  ^25 liters per minute (0.8 - 1.0 CFM)  rated flow
 capacities.  Generally speaking, in the meter box calibration procedure,
 this designed How rate is routinely exceeded.  There are two
 considerations  in  this fact.   The first is that the  wet test meter in
 all probability loses  some of its accuracy at flow  rates  exceeding the
 design value.  This fact would  tend to skew  the elevated  flow data
 points used in the least squares calculation.
     Secondly, in using these smaller  flowmeters (a  use dictated  by
 economics) the procedure develops pressure differentials  across the wet
 test meter which are not recognized to be significant  in  the  EPA
 publications on this subject.   All  of  the EPA  documentation  assumes that
 the pressure drop across  the  wet test  meter  is  essentially  equal to
 zero.  In these small test  meters being used  as  described,  this pressure
 drop assumes significant  magnitudes, on the order of 2 to 3 inches  of
water,  and therefore, should  be included  in the  computations  associated
with the meter box calibration  in the  same sense that the pressure  drop
across  the orifice is included  where its  magnitude perhaps is higher but
not so  much so  that one could justify  ignoring equivalent pressure  drops
in the  wet test meter.  This  erroneous  assumption with respect  to wet
                                   24

-------
  test meter pressure  drop  also  contributes to some inaccuracy in the
  calculated data  of a meter box calibration.  It is reasonable to think
  that these additional built-in errors could adversely affect the slope
  intercept  by  least squares computation.
            Response:  Exceeding the capacity of the wet test meter will
  certainly  affect the accuracy of the calibration.   The October 10,  1982
  (47  FR 44350) proposal  addresses the problem to some extent by  allowing
  the  use and specifying a procedure for calibrations  using  a dry  gas
 meter as the standard.   The standard dry gas meter  is calibrated against
 a wet test meter or spirometer  periodically.  Otherwise, this issue is
 not within the scope  of  the QC.
      16-6   Comment:   As  indicated  above,  I have made  a limited survey
 of emission testing groups  in our  local  area to determine whether the
 ±1.5  percent QA check as proposed would  be realistic.  It is quite
 evident  that some among this group can comply with the ±1.5 percent.
 Others cannot.
         Response:  See Response 16-4.
      16-7   Comment:   The proposed QA check does  not  suggest that  the
 coordinates  of zero and zero be included in the  series of data  points
 used  to determine the least squares fit equation.  The case may be
 argued that  this point is part of the calibration  procedure and should
 be included.  It is  my suggestion  that this  point be  included.  It was
pointed out earlier  in this text that a 2 percent change  in  one
coordinate resulted  in an  approximate 1/2 percent change  in  the
                                   25

-------
predicted SCFM.  The inclusion of the origin or the coordinates  zero  and
zero in the set of data materially reduces  this problem.
          Response:  The development of a  linear,  best-fit  curve with
calibration data is no longer required.
     16-8  Comment:  In conclusion,  my recommendations would  be  that EPA
consider the alternate and more simple QA  check described herein, and
secondly, that if EPA feels it necessary to adopt  the more  complex
procedure proposed, that they consider making  the  compliance  limits more
liberal, say +3.0 percent in order that the field  personnel have a
better chance to comply in view of potential  limitations in the  least
square computation.
          Response:  We agree.   The  promulgated QC  check is a one-point
check with a recommended +3 percent  tolerance.
Commenter IV-D-17
     17-1  Comment:  The equation for calculating  Q^ in Section  4.4,
Orifice Factor Determination,  is incorrect  and  several of the variables
have not been correctly defined.
     vm(std) should be calculated according to  Equation 5-1 of Method 5.
          Response:  See Response 2-1.
     17-2  Comment:  The equation for OF, does  not correct the OF for
pressure and temperature.  Since Qj  is calculated  under standard
conditions, OF should also be calculated for standard conditions.
                                   26

-------
            Response:  The OF factor equation is no longer part of the
 QC check.
      17-3  Comment:  It has been specified that the agreement between
 the estimated and the calculated values of Q should be within 1.5
 percent, at or near AH@.  However,  the following error analysis
 suggests this requirement to be too restrictive.
      Some calculations based on orifice calibration  data  provided to
 the National  Council  of the Paper Industry for Air and Stream
 Improvements,  Inc.  by member companies  result  in a  total  error of
 +5 percent.
      It is  our  suggestion that  since  error  analysis has shown much
 greater possible  errors  than are  allowed in the present proposal, the
 allowed error fn  the  correlation  equation should be based upon actual
 field data.
          Response:   See Response 16-4.
     17-4   Comment:   Consideration should be given to making these
 requirements optional with a higher number (at least 5 percent)  for
 percent  difference between actual  and estimated values.  The present
 Method 4 and 5 requirements  of post-test meter check  and 5 percent
 agreement are adequate to satisfy  the QA needs.
          Response:   See Response  7-1.
Commenter IV-D-18
     18-1  Comment:   These proposed  changes  are  not justified  by a
hard data base,  but  only  by  the  unsupported  statement  that "(t)he
                                   27

-------
 current  regulation involves only limited QA requirements and, as a
 result of this proposed regulation, the quality of compliance data will
 improve."  As before, we urge EPA to withdraw this proposal  pending
 the presentation of hard supporting data and the public's opportunity
 to review and comment on the supporting data base.
          Response:  The Agency has a mandate to include in  all  test
 methods QA and QC procedures to assure and assess the accuracy of
 measured data.  The Agency feels the promulgated QA procedures are
 adequately supported and are achievable.
     18-2  Comment!  The term (- AH/13.6)  in the first equation  should
 carry a positive rather than a negative sign.
          Response:  See Response 2-1.
     18-3  Comment:  The equation will  fail  for English  units if the
metric constant of "273" is used,  but this is  not made clear.
          Response:  We agree.
     18-4  Comment:  The parameter Tm must carry °C  units whereas  Tm as
defined in Section 6  can carry either °K or °R units  but not  °C  units.
          Response:  Agree.
     18-5  Comment:  If Tm is  to carry  °C  units,  then the constant "273"
should be carried to  four significant figures  or 273.2°  for consistent
precision.
          Response:  Tm is defined in °R.
     18-6  Comment:  The constant 0.3855 should be 0.3858 as  already
defined in Section 6.   The 0.3855  value results  when  293°K, rather than
                                    28

-------
  293.2°K,  is  divided  by  760.0, mm Hg, but four-place precision demands the
  more  precise 293.2°K value.
           Response:  This constant does not appear in the promulgation
  version.
       18-7  Comment:  Although it is not clear from the definition given
  for Qi, it reflects volume flow rate at standard,  not actual,
 conditions.
           Response:  This is correct.
      18-8  Comment:  The nomenclature  of OF,  or orifice  factor, does
 not seem appropriate  for a grouping  of variables and  tends to be  easily
 confused with the more conventional meaning of "orifice factor," which
 is  an  orifice meter constant.
           Response:   The conventional  terminology in defining the
 constant for  an orifice  is "orifice coefficient."
     18-9  Comment:  There is also a fundamental  problem with the
 consistency check procedure of Section 4.4.3.   It  calls for the  dry gas
 meter  flow at standard conditions to be compared with  the orifice  flow
£t actual condrtions.   This problem would  be eliminated by  inversion of
 the (T0/P) term to give (Pb/T0)  within  the  OF  equation.   This simple
 rearrangement can  be shown to yield an  orifice flow rate  at standard
conditions.
          Response:  See  Response 17-2.
                                   29

-------
     18-10  Comment:   One final  problem with  Section 4.4.3  is  the use of
the parameter AH@.  This parameter is  not  presently defined in Method 5.
          Response:   This is defined in the proposed revisions.
Commenter IV-0-19
     19-1  Comment:   The proposed changes  are good and  have, in  fact,
been used by the American Boiler Manufacturers  Association  members
successfully for about 10 years.   Not  only does the proposed change
result in a better quality test,  but it also  makes the  test easier to
perform.  To insure greater QA  results should be compared with a
theoretical  combustion analysis  and should be within +5 percent
(inclusive of moisture in air!),  and Method 5 results should be checked
for percent isokinetic sampling  for each sample point.  This proposed
change for Method 5  would be tedious.   However,  as it now stands, test
result averages may  indicate an  isokinetic rate within  the  required 90
percent to 110 percent when none of the individual test points were
within this range.
          Response:   Such additional QC or QA checks are not within
the scope of this proposal.
Commenter IV-D-20
     20-1  Comment:   We believe the concept of  field checking  the
volume metering system is a good idea; it  could avoid costly retests.
We do not, however,  feel that the proposed agreement limit  of  1.5 percent
is realistic.  Potential reading errors in the  orifice  pressure,
barometric pressure,  and temperature can account for nearly all  of the
                                    30

-------
 1.5 percent variation with no actual change in the dry gas meter.
      The meter should be checked in the field before and after  the
 testing program to assure accurate data during the test.   Predicted  vs.
 measured values should agree with +5 percent,  as  the method currently
 allows.  It seems completely unreasonable to  insist on +U5 percent
 agreement under field conditions when it has  been  generally recognized
 that +5 percent variation in the before and after  calibration is
 acceptable.
           Response:   See Responses  16-4  and 16-8.
      20-2  Comment:-   The second  listing  of Section 4.4 appears to be
 improperly labeled, and  should be corrected to read Section 4.4.1.
          Response:   See Response 1-1.
     20-3  Comment:   In  Section 4.4.1 (corrected reference) the  units
 for  AH.,-  should  be mm  H20  (in. ^0).
          Response:   See Response 1-2.
     20-4  Comment:   In  Section 4.4.1 (corrected reference) the  term
 11V should be  defined.  We suggest the same definition used in  Section
 6.1  for this term.
          Response:  See Response 1-2.
Commenter IV-D-21
     21-1  Comment:   The main point of confusion with  Methods 4  and 5
are timing and applicability.   When  specifically is  "prior  to the start
of the first  emission  test run of a  series of  runs"?   If a  boiler is to
be tested on  one day  for  velocity,  reference moisture,  and NOX, then
                                   31

-------
next day particulates, would one meter field QA  check  suffice on the
first day?  Would the meter field QA check  be  acceptable on the
afternoon prior to the first day of testing?   If tandem boilers are to
be tested consecutively with the same console, would one field check
suffice?
          Response:   The QC check is recommended prior to the first
test in order to check the operation of  the metering system.  Repeat
checks are not necessary as other QA/QC  checks are required in the
method, e.g, leak checks, post-test calibrations.  See also Response
5-1.
     21-2  Comment:   The QA procedure makes no mention of alternatives
when in the field, other than  recalibration, a laboratory operation.  If
the meter does hot meet specification or if the  calibration paper is
lost, the test crew would have no alternative  but to obtain another
specially calibrated meter console.   Even if a replacement (normally)
calibrated dry gas meter were  obtainable near  the field site, it would
be unusable due to the lack of this special extra calibration.
          Responses:   See Response 5-1.
     21-3  Comment:   The calculation equation  and legend contain some
omissions and questionable material.   Y  factor is not. stated to be
individual or average.  There  is no Tm figure  in the legend, and the
addition of 273 is for °K only;  no mention  is  made of conversion to °R.
AH-f is not listed as primarily having values in  mm Hg.  Also, why is
                  i
the AHj factor subtracted?
                                    32

-------
            Response:   See Responses  1-1, 1-2, 1-3.
       21.4  Comment:   Upon review of the existing QA requirements and
  recommendations, another calibration seems primarily redundant.  A sample
  meter console is required to be fully calibrated initially,  with
  mandatory  calibration checks to be performed after each field use.
  These calibration checks may also be used as pretest checks  for the
  following  sample program.   The QA objectives seem substantially
  imbalanced when it is entirely optional  to perform pretest leak checks
 of the sample system (post-test checks,  of course,  mandatory),  but  a
 specific extra  pretest calibration  at a  specific  time is mandatory.
 Obviously,  it is  in  the best interest of  the  sampler (and his client) to
 insure that the  equipment will  perform adequately.  However,  it js the
 sampler's  ultimate responsibility to obtain a valid sample,  with QA data
 proving  its validity.   How the  equipment is maintained and calibrated
 before the  test is a  responsibility belonging solely to the sampler.
      If  there is a need  for  additional  QA, it is the enforcement of
 current  (and  valid) QA procedures.  If not enforced, additional  QA
 procedures  will  only hinder  the personnel  responsible enough  to  perform
 and present the required QA work.
          Response:   See Response 5-1 and 7-1.
Commenter IV-D-22
     22-1  Comment:   Section  4.4.3.   This  field  check is a good  QA tool
but requires additional  time  to  perform.  The  agreement limit of +1.5
percent between  the calibration  in the laboratory  and the field  check is
                                   33

-------
tight, and there may be difficulty in meeting it.   The  value  of  this
check is to ascertain that no changes have  occurred in  handling
equipment.  Therefore,  if the field check is  performed  and  the limits
are met, there is little value in continuing  the now  required post
calibration.  If the field check is successful  and  the  post calibration
fails, the change most likely would be caused by post-testing handling.
A post-test field check would make more sense.
          Response:   See Response 5-1,
     22-2  Comment:   Methods 4 and 5,  in most instances, pre- and
post-test calibration fall  within acceptable  limits.  A field check of
the volume metering system's calibration may,  on rare occasions, identify
a problem incurred during handling after calibration.   However, the
company takes'exception to the +1.5 percent tolerance which is the same
as laboratory limit and which may be difficult  to attain during adverse
weather conditions.   A  field check tolerance  of +5  percent  limit is more
realistic.
          Response:   See Response 16-4.
                                    34

-------
                       Table 1.   LIST OF COMMENTERS
Document Number
                                                Commenter
  IV-0-1
  IV-D-2
  IV-0-3
  IV-0-4
 IV-0-5
 IV-0-6
 IV-D-7
 From Jon M. Rueck,  Kansas Department
 of Health and Environment to Docket
 Number A-82-04.   Subject:   Comments
 on proposal of revisions  to Methods 4
 and 5.

 From Jim Steiner, Pape and Steiner
 Environmental  Services to Roger
 Shigehara,  U.S.  Environmental
 Protection  Agency.   Subject:
 Comments on proposal  of revisions to
 Methods 4 and  5.

 From C.D. Malloch, Director  of
 Regulatory  Management,  Monsanto
 Company to  Docket Number A-82-04.
 Subject:  Comments on  proposal of
 revisions to Methods 4  and 5.

 From Jerry  Wayne Powell, Texasgulf
 Chemicals Company to Docket Number
 A-82-04.  Subject:  Comments on
 proposal  of revisions  to Methods 4
 and  5.

 From Robert E. Wooten, North Carolina
 Department  of Natural Resources and
 Community Development to Docket
 Number A-82-04.  Subject:   Comments
 on proposal  of revisions to
 Methods 4 and 5.

 From Frank Rower, Armco, Inc. to
 Docket Number A-82-04.  Subject:
 Comments on proposal  of revisions
 to Methods 4 and 5.

From Dale S. Harmon,  Texas Industries,
 Inc. to Docket Numbers A-82-05  and
A-82-04.  Subject:   Comments on
proposals of revisions .to  Methods 3,
4, and 5.
                                  35

-------
                      Table 1.   LIST OF COMMENTERS
                               (Continued)
Document Number
               Commenter
  IV-0-8
  IV-D-9
  IV-0-10
  IV-0-11
  IV-0-12
  IV-0-13
From James H.  Boyd,  Director of
Environmental  Affairs,  Newmont
Services Limited to  Docket Number
A-82-04.  Subject:   Comments on
proposal of revisions  to Methods  4
and 5.

From Michele E.  Eller,  Aluminum
Company of America  to  Docket Number
A-82-04.  Subject:   Comments on
proposal of revisions  to Methods  4
and 5.

From J.F. Kane,  Plant  Manager,  E.I.
duPont de Nemours and  Company to
Docket Number A-82-04.   Subject:
Comments on proposal of revisions to
Methods 4 and 5.

From W.J. Barrow., Manager,
Environmental  Remitting and
Programs, Florida Power and Light
Company to Docket Number A-82-04.
Subject:  Comments  on  proposal  of
revisions to Methods 4 and 5.

From Bruce G.  Hawks, Supervisor,
Quality Assurance,  Entropy
Environmentalists,  Inc. to Docket
Number A-82-04.   Subject:  Comments
on proposal of revisions to Methods 4
and 5.

From Danny E. Sjolseth, Section
Manager, Air Technology Research  and
Development, Weyerhauser Company  to
Docket Numbers A-82-05 and A-82-04.
Subject:  Comments on  proposals of
revisions to Methods 3,4, and 5.
                                    36

-------
                       Table 1.   LIST OF COMMENTERS
                               (Continued)
Document Number
               Commenter
  IV-D-14
  IV-0-15
  IV-0-16
  IV-D-17
  IY-D-18
From L. Blaine DeHaven, Chief, Source
Testing and Monitoring Section,
Pennsylvania Department of
Environmental Resources to Docket
Number A-82-04.  Subject:  Comments
on proposal of revisions to Methods 4
and 5.

From Harold E. Hodges, Director,
Tennessee Air Pollution Control
Division, Tennessee Department of
Public Health to Docket Numbers
A-82-04 and A-82-05.  Subject:
Comments on proposal of revisions to
Methods 3, 4, and 5.

From Robert S. Sholtes, Sholtes and
Koogler Environmental Associates  to
Docket Number A-82-06 (changed to
A-82-04).  Subject:  Comments on
proposal of revisions to Methods  4
and 5.

From Ashok K. Jain, Engineering
Projects Research Manager,  National
Council for Air and Stream
Improvement, Inc. to Docket Number
A-82-05 (added to A-82-04).   Subject:
Comments on proposal of revisions to
Methods 4 and 5.

From S.I.  Estes,  Union Camp  Corp. to
Roger Shigehara,  Emission Standards
and Engineering Division.   Subject:
Comments on proposals of revisions to
Methods 3, 4, and 5.
                                   37

-------
                       Table 1.   LIST OF COMMENTERS
                               (Continued)
Document Number
               Corranenter
  IV-D-19
  IY-0-20
  IV-0-21
  IV-D-22
From W.H.  Axtman,  Executive Director,
American Boiler Manufacturers
Association to Roger Shigehara,
Emission Standards and Engineering
Division.   Subject:   Comments  on
proposals of revisions to Methods  3,
4, and 5.

From Richard H. Russell,  Supervisor,
Environmental  Testing, Owens-Illinois
to Docket Number A-82-05  (added  to
A-82-04).   Subject:   Comments  on
proposals of revisions to Methods  3,
4, and 5.

From Randall J. Richert,  Southwestern
Laboratories to Docket Number  A-82-Q5
(added to A-82-04).   Subject:
Comments on proposal of revisions  to
Methods 3,4,  and 5.

From R.A. Reckert, Vice President
Fossil and Hydro Production,
Northeast Utilities to Docket  Number
A-82-05 (added to A-82-04). Subject:
Comments on proposal of revisions  to
Methods 3, 4,  and 5.	
                                    38

-------
                                      TECHNICAL REPORT DATA
                              (Please read Instructions on the reverse before completing)
  REPORT NO.
  EPA 450/3-83-020
                                2.
                                                                3. RECIPIENT'S ACCESSION NO.
4. TITLE AND SUBTITLE
  Revisions to  Methods 4 and  5,  Appendix  A of 40 CFR
  Part 60 - Summary of Comments  and Responses
5. REPORT DATE
 December, 1983
                                                                6. PERFORMING ORGANIZATION CODE
  AUTHOR(S)
  Emission Measurement Branch
  Emission Standards and Engineering Division
8. PERFORMING ORGANIZATION REPORT NO.
                           VIE AND ADDRESS
  U.S. Environmental  Protection  Agency
  Mail Drop 19
  Research Triangle Park, N.C.  27711
10. PROGRAM ELEMENT NO.



11, CONTRACT/GRANT NO,
           IG AGENCY NAME AND ADDRESS
  U.S.  Environmental  Protection Agency
  Mail  Drop 19
  Research Triangle Park, N.C.  27711
                                                                13. TYPE OF REPORT AND PERIOD COVERED
14, SPONSORING AGENCY CODE
lu ABSTRACT
       Document is summary  of comments  submitted as a  result of the  September 7,  1983,

  proposal of  revisions to  Methods 4 and  5  to add quality assurance  and quality control

  procedures.   Included are the Agency's  responses to  the comments.
                                  KEY WORDS AND DOCUMENT ANALYSIS
                   DESCRIPTORS
tl DISTRIBUTION STATEMENT
                                                 b.lDENTIFlERS/OPEN ENDED TERMS
                                                 19. SECURITY CLASS (This Report)'
                                                 20. SECURITY CLASS (This page)
                                                                              c.  COSAT1 Field/Group
               21. NO. OF PAGES
              22. PRICE

-------