United States
             Environmental Protection
             Agency
             Region 5
            Environmental Monitoring Branch EPA-905'4-82-002
            230 South Dearborn Street   September, 1982
            Chicago, Illinois 60604
vvEPA
Non-Methane
Organic Compound
Continuous Monitoring in
Cleveland and Cincinnati
1981 Ozone Monitoring Study
PEDCo Environmental, Inc.
11499 Chester Road
Cincinnati, Ohio 45246-0100
Report No. EPA-905/4-82-002
Contract No. 68-02-3512
Work Assignment No. 22
PN 3525-22

-------
    NON-METHANE ORGANIC COMPOUND
       CONTINUOUS MONITORING
    IN CLEVELAND AND CONCINNATI:
     1981 OZONE MONITORING STUDY
             Prepared by

      PEDCo Environmental, Inc.
         11499 Chester Road
    Cincinnati, Ohio  45246-0100
     Report No.  EPA-905/4-82-002
       Contract No.  68-02-3512
       Work Assignment No.  22
             PN 3525-22
             Prepared by

U.S.  ENVIRONMENTAL PROTECTION AGENCY
       536 SOUTH CLARK STREET
      CHICAGO, ILLINOIS  60605
            August, 1982

-------
                                  CONTENTS
List of Figures                                                      iii
List of Tables                                                       iii

1.0  Introduction                                                    1-1

2.0  Calibration Procedures                                          2-1
     2.1  Introduction                                               2-1
     2.2  NMOC Calibration                                           2-1

3.0  Quality Assurance Program                                       3-1
     3.1  Introduction                                               3-1
     3.2  Zero and Span Checks                                       3-1
     3.3  Precision Checks                                           3-4
     3.4  Systems Audit                                              3-4
     3.5  Data Validation and Audit                                  3-4

4.0  Site Operation Summary                                          4-1
     4.1  Rickoff School Site                                        4-1
     4.2  St. Vincent's Site                                         4-1
     4.3  University CAM Site                                        4-1
     4.4  Norwood Site                                               4-2

APPENDIX A     Local Operators Standard Operating Procedure (SOP)     A-l

APPENDIX B     Precision Check Program                               B-l
                                     ii

-------
                                                                                         V
                                   FIGURES


Number                                                                Page

  1       Calibration Data Form                                       2-3

  2       Zero Span Data Sheet                                        3-2
                                   TABLES
Number                                                                Page

  1       NMOC Calibration Gas Concentrations as Certified by
           U.S.  EPA                                                   2-1

  2       NMOC Audit Summary                                          3-5
                                     iii

-------
                                                                      in1.111: mi  ir mi  iir
                                 SECTION 1.0
                                INTRODUCTION

     Revisions to the State Implementation Plan (SIP) for ozone are required
in 1982.  The  State  of Ohio must obtain Non-methane Organic Compound (NMOC)
data for days  of high ozone concentrations during the summer of 1981.   Data
were especially  needed  in the vicinity of Cleveland  and Cincinnati  to pro-
vide the needed  input into the predictive model that will be used to demon-
strate attainment of the ambient ozone standard.
     To obtain this  data two NMOC analyzers were operated in both Cleveland
and  Cincinnati  by  the  local  agency.  PEDCo  Environmental, Inc. was  con-
tracted to  assist the  local  agencies in the  installation,  start up,  main-
tenance and quality assurance of the four analyzers.
     The sites  located  in  Cleveland were the  Rickoff School  site at 147th
Street, and  the  St.  Vincent's site  located  at Woodland  and 22nd Streets.
The  Cincinnati  sites were  the University CAM site located  at  Vine  and St.
Clair Streets, and the Norwood site located at 300 Harris Avenue.
     Instruction was  given  to the local operators  on the routine operation
and maintenance of the NMOC analyzers, according to the procedures published
in  the  EMSL  report  EPA-600/4-81-015,  March  1981.   Standard  Operating
Procedures (SOP)  (Appendix A), were supplied to the local operators.
     The following  sections describe the calibration procedure and quality
assurance activities.
                                      1-1

-------
                                 SECTION 2.0

                           CALIBRATION PROCEDURES


2.1  INTRODUCTION

     The NMOC analyzers were  scheduled to be calibrated approximately every

two weeks,  however,  this  schedule  was not always met.  This was  primarily

due to instrument malfunctions, and scheduling difficulties.

     Every effort was made to calibrate each site within the prescribed time

frame.  However, priority  was  given to those instruments which  were subject

to downtime or drifting out of calibration.  Those instruments which did not

exhibit  excessive  drift  were given  a  lower  priority and  were  calibrated

after the problem instruments had been serviced and calibrated.


2.2  NMOC CALIBRATION

2.2.1  Calibration Equipment

     The  NMOC analyzers  were calibrated  using  EPA  certified methane  and

propane calibration  gases.   The  gases were certified prior  to  start of the

project.  Table 1 lists the tanks and certified concentrations.


                TABLE 1.   NMOC CALIBRATION GAS CONCENTRATIONS
                          AS CERTIFIED BY U.S. EPA
Cyl
Bal
Bal
Bal
Bal
. no.
1236
1201
1222
1205
Concentration
746 ppm CH4
748 ppm CH4
249 ppm C3H8
250 ppm C3H8
                                      2-1

-------

These gases  were then  diluted  with compressed  air  (<0.01 ppm HC)  using  a

porous plug  dilution system.   The  system used  throughout the project  was

Serial No.  RTI-PE-103.   Flows  were  measured with a Hastings  NBS  traceable

bubblemeter.

2.2.2  Calibration Procedure

     The following is the  stepwise  procedure used to calibrate the MSA 11-2

continuous analyzer.   The analyzer should be allowed  to warm up for approxi-

mately 24 hours prior to calibration.

     NOTE:   The  analyzer  was  calibrated and  operated with the  elec-
     trometer in the subtraction mode.

All  calibration  data was  recorded  on a  calibration data  form  (Figure 1).

The calibration data has been sent to the Project Officer.

     1.    If  this  is  an  initial calibration,  shut  off  the vent  flow for
          checking;  it  should be checked and adjusted prior to the calibra-
          tion.

     2.    Check  the  moisture drop-out traps  located  on  the  back of  the
          analyzer for  the accumulation  of water.   Drain  any  water accumu-
          lation in  the second  water  trap  toward the instrument.  In the
          first  water   trap  after the  sample  pump,  maintain  % inch  to  2
          inches of water.   Any water accumulation higher than % inch should
          be drained to that level.

     3.    Quickly check the back-pressure regulator vent flow located inside
          the  oven.   Vent  flow from  each  regulator  should  be 400 to 600
          cc/min.  These flows should be closely matched.

     4.    Check analyzer sample  inlet  pressures.   The air pressure gauge on
          the front  of  the instrument should be  20  ±  1 psig.   The pressure
          gauge  between the water  traps  and the instrument on  the back of
          the  analyzer  should be 30 ± 2 psig.   The pressure  gauge between
          the two water traps  should be greater  than  40  psig.   If the pump
          will not maintain  at  least 40 psig,  check for  leaks in the water
          trap system or water and dirt in the sample pump.

     5.    Check that the range  selector  switch is in the 0 to 10 ppm range.

     6.    Disconnect the  analyzer's sample  line from  the ambient manifold
          and  reconnect it to  the  calibration  system's manifold.   Cap the               f
          port on the ambient manifold and any open ports on the calibration               I
          system's sample manifold.


                                      2-2

-------
                          CLEVELAND/CINCINNATI
                 NONMETHANE ORGANIC COMPOUND MONITORING
                               PN 3525-22

 CALIBRATION  DATA FORK FOR MSA MODtL 11-2 HYDROCARBON ANALYZER
 Site: 	     Snubber RT1-PE r
 Date:                                   Bubble Meter * _
 Analyzer  S/N.                           Zero Air Cyl •
Analyzer off  line at: _     CH$ Cyl * _  _ ppm
Analyzer on line at: _     C3HB Cyl • _  _ ppm
NOTES: _     Temperature _ "C
_     Barometric Pressure _ " Hg


Analyzer Pot  Settings:  Zero NMOC
                                  _
                        Span NMOC _ CH4
Correction Factor for Temperature, Barometric Pressure, and Vapor Pressure of
Water:


                     ^ x -.	r  *	Correction factor
           29.92         (

METHANE CALIBRATION.

Dilution floK-. 	psi      CH  Zero	%         NKOC Zero


/ 10°°  > x (      ) • 	 SCCK Zero Air
«1 Point:  Pollutant pressure: _ psi Toggles 1   2   3
          x (      ) • _ SCCK CH4
                	'. Chart CH4 	V Chart NMOC

«2 Point:  Pollutant pressure:   	 psi Toggles 1   2   3


7—^	r  X (       ) * 	 SCCM CH,
                 _ '. Chart C«4 _ i Chart NMOC

13 Point: Pollutant pressure:         _ psi Toggles 1   23
j
               (        ) • _ SCCM CH
                             1 Chart CH4 _ '. Chart NT10C
                             Figure  1.
                                 2-3

-------
   •4 Point:  Pollutant pressure:   	 psi Toggles 1   2   3
   T	r * (     ) • 	  SCCK CH.
                             ppm CH.
                             % Chart  CH.             % Chart NMOC
        CHj SLOPE: 	     INTERCEPT:
   NMOC CALIBRATION:
   Dilution flow: 	 psi       CH^  Zero 	X          NMOC Zero
    10DD     ,     v
                                  SCCM  Zero Air
      Point:  Pollutant pressure:  	 psi Toggles 1   2   3
      10     - {     )  •	 SCCM  c,H,
   I      T   v
            	^ - 	 PP"1 C3Hg
                     	 % Chart CH4 	 % Chart NMOC
   «2 Point:  Pollutant pressure:  .	psi Toggles 1   2   3
      10      - (      ) -  	 SCCM  C,HC
            T
            ^
                     _ I Chart  CH4 _ % Chart NMOC
   • 3 Point:  Pollutant pressure:  _ psi Toggles 1   2   3
      10
   •T
                (      )  -  _ SCCM  C,HS
                                  SCCM C,HD
                             8 Chart CH^ _ S Chart NMOC
        NMOC SLOPE:  _    INTERCEPT: _       CORR:
Figure  1  (continued)
                                      2-4
                                                                               il!
                     _ I Chart  CH4 _ * Chart NMOC                                             j?
      Point:  Pollutant pressure:  _ psi Toggles 123                                          :

-------
                                                                 Ill 1 .  XI
                                                                                     P:

 7.    Supply an  atmosphere of  the  zero standard  to the manifold  at a             te
      flow rate  that is  20  to 50%  greater than  the  analyzer's  sample
      flow demand.   DO  NOT PRESSURIZE THE ANALYZER'S SAMPLE INLET.   The
      test atmosphere must contain  an ambient level of  oxygen  and  must
      not contain more than 0.1 ppmC TOC.

 8.    Adjust  the  analyzer's  zero  controls  for  both  channels to  the
      desired baseline responses.   A 5% of full scale positive offset on
      the recording device is recommended to observe any negative drift.
      Use either  the analyzer's zero pots or the recorder's controls to
      obtain the  offset.   Ensure  that the  responses from  both  channels
      are .equal  before  recording  the responses on  the  calibration  data
      sheet.

 9.    Supply an atmosphere of methane standard to the calibration mani-
      fold at a  flow rate that is 20 to 50% greater than the analyzer's
      sample flow demand.  The methane concentration  should  be between
      70 and 90% full scale.

10.    Adjust span pot #2  (which  controls CH4 response  from  FID  #2) to
      provide the analyzer response calculated as follows:

      D      ~ -  Sample concentration u   ,  ^ ,     .,  .
      Response =  	c	rj^-	 X url  + Zero offset

      where

      Response = Response of the recording device measuring the
                 analyzer output in recording device units.

      Sample concentration = Concentration of the calibration standard
                             delivered to the analyzer in ppmC.

      URL = The upper range limit of the analyzer in ppmC.

      url = The upper range limit of the recording device in recording
            device units.

      Zero offset = The amount the recording device response is set
                    above the zero baseline while the analyzer is
                    measuring the zero calibration standard (in
                    recording device units).

11.    If  step   10 results in  a  span pot setting  greater than  300,
      decrease  pressure  to   H2  gauge  #2  until  the  CH4   response  has
      increased to  a point where the span pot can  be  reduced  to about
      250.  If the span pot  setting is less than 200, increase pressure
      to  H2  gauge #2 until  the CH4  response has  decreased  to a point
      where the span pot can be increased  to  about 250.  DO NOT EXCEED
      9.0 psig H2.

           NOTE:    IF THE  FID 'RESPONSE  INCREASES  WHEN INCREASING
           HYDROGEN  PRESSURE, THE BURNER  IS OPERATING ON THE WRONG


                                  2-5

-------
           SIDE OF  ITS PEAK  RESPONSE  CURVE.  CONTINUE  INCREASING
           HYDROGEN PRESSURE UNTIL RESPONSE  STARTS  TO DECREASE AND
           THE CORRECT SPAN POT SETTING IS OBTAINED.

12.    Set span  pot #1 NMOC to the  same dial setting  (not  response)  as
      span pot #2.

13.    Repeat steps 7,  8,  9,  10, 11,  and 12 if span  pot adjustments were
      necessary or if hydrogen pressure gauge #2 was adjusted.

14.    Continue  sampling   the   methane  standard.    For  the   FIDs  to  be
      balanced, the  NMOC  response  must  be  within  ±1.0%  of  the  zero
      response obtained in step 8.   If the FIDs are  balanced, go to step
      16.

15.    Balance detectors:   If the FIDs are not balanced:

      (a)  Check that  both span pots  are dialed to the same setting.

      (b)  Locate  the  range  change  board  (inside  the  electrometer
           assembly located in the oven) and rotate  the balance controls
           of  channels  1  and  2 (accessible through  holes  in  the elec-
           trometer  cover)  to  their  maximum clockwise  position.   Make
           adjustments quickly to minimize heat loss!

      (c)  Record  both   hydrogen   pressure   gauge   readings.   Adjust
           hydrogen  pressure   gauge #1 until the NMOC response  to the
           methane standard equals the earlier NMOC  response to the zero
           standard (step 8).

           Do not adjust  the  hydrogen pressure outside  the  5  to 9 psig
           range.   Increase hydrogen pressure to decrease NMOC response,
           or vice versa.   See the note in step 11.

      (d)  If balance is achieved,  repeat steps 7 through 15.

      (e)  If  balance  is not achieved  by  step  c,  adjust  the  NMOC
           hydrogen pressure  (gauge  #1)  or  both hydrogen  pressure,  if
           necessary, to get as close to balance (NMOC response within ±
           1% of NMOC  zero  response)  as possible.   Do not  exceed the 5
           to  9 psig hydrogen pressure range.  Then,  using the balance
           controls on the  range change board located inside  the elec-
           trometer  (accessible  through   holes   in  the  electrometer
           cover),  rotate  the  channel  #2  balance  control  counter-
           clockwise  to  increase  the  NMOC  response,   or  rotate  the
           channel #1  balance  control  counter-clockwise to decrease the
           NMOC response.   Repeat steps 7 through 14.

16.    After  the FIDs  are  balanced,  record responses on the Calibration
      Data Sheet.  Determine  the CH4 channel's response to three addi-
      tional  concentrations  of the  methane standard  that are  spaced
      approximately equally over the analyzer range.  (The NMOC
                                  2-6

-------
                                                                                    iii
                                                                                    ""   N,
                                                                                          I
          channel's response should  remain  equal  to the earlier response to
          zero standard.)  Record the CH4 and NMOC channel's responses (from
          the  recording  device)  on the  data  sheet.   Using a  calculator,
          perform  a  least squares  linear regression  corresponding  calibra-
          tion concentrations.   The  calibration concentrations  should be in
          units of  ppmC  and should be entered  into  the calculator  as  the
          independent variable  X.   The CH4  channel's response should  be in
          units  of  the   recording  device  and  should   be  entered  as  the
          dependent variable Y.   A  correlation coefficient (r)  of 0.9996 or
          better verifies that the CH4 response is linear.   (If  the response
          is not  linear, plot  the  data  and determine if an error  has been
          made in  data entry or in determination  of  calibration concentra-
          tion.)   Obtain  the   slope  and intercept  of  the regression  and
          record the equation in the following form:
          CH4 Response =  CH4 Slope x Methane concentration + CH4 Intercept

          where

          CH4 Response =  Analyzer's CH4 channel reading  in recording device
                         units (see note following step  10).

          CH4 Slope = Regression slope in recording device units per ppmC.

          Methane concentration = Calibrated methane concentration in ppmC.

          CH4 Intercept = Regression intercept in recording device units.

Post  the  CH4  channel's  multipoint  calibration  curve  equation  on  the

analyzer's recording device and also on the  Calibration  Data Sheet.

    17.    Supply an atmosphere  of propane standard to the calibration mani-
          fold at  a  flow rate  that is 20 to 50% greater than the analyzer's
          sample flow  demand.   The propane concentration  should be between
          70 and 90% of full scale.

    18.    Adjust span pot  #1 (which controls NMOC response  from FID #1) to
          provide the desired analyzer response.

    19.    If an  adjustment  is  made in step 18, recheck  the  NMOC channel's
          response to the zero standard and  adjust zero  pot #1 if necessary.
          Record the stable  zero  responses  from the NMOC  and CH4 channels.
          Sample the propane standard, and again record  the stable responses
          to propane.   (CH4 response should  be equal to  the earlier response
          to zero air.)

    20.    Determine the  NMOC channel's response to three additional concen-
          trations  of  the  propane  standard  that are  spaced  approximately
          equally  over  the  analyzer  range.    (The  CH4 channel's  response
          should  remain  equal  to the earlier  response to  zero standard.)
          Record the  NMOC and  CH4  channel's  responses  (from the recording
          device)  on  the  Calibration  Data  Sheet.   Using  a  calculator,
          perform  a  least squares  linear regression of  the  NMOC channel's

                                      2-7

-------
      response  (to  propane  and zero  standards)  and the  corresponding              |*
      propane  calibration  concentrations.   The  calibration  concentra-              |
      tions  should  be in  units of ppmC  and should  be  entered as  the              fe
      dependent variable Y.  A  correlation  coefficient  (r) of 0.9996 or              I
      better  verifies  that  the  NMOC  response  is  linear.   (If  the              f
      response  is  not linear,  plot the  data and determine  if  an  error              *•
      has been  made in data entry or  in determination of  calibration
      concentration.)  Obtain the  slope  and intercept of the regression              I"
      and record the equation in the following form:                                   f

      NMOC Response = NMOC Slope x NMOC Concentration +  NMOC Intercept                t

      where                                                                           ;

      NMOC Response = Analyzer's NMOC  channel reading in recording
                      device units (see note following step 10).

      NMOC Slope = Regression slope in recording  device  units per ppmC.                I

      NMOC Concentration = Calibrated  NMOC concentration in ppmC.                      j-

      NMOC Intercept = Regression intercept in recording device units.                 ;•

      Post the NMOC channel's multipoint calibration curve equation
      on the analyzer's recording device and on the  Calibration Data                  I
      Sheet.                                                                          u

21.    Disconnect the analyzer's sample line from  the calibration system's
      manifold and reconnect it to the ambient manifold.                               I

22.    Be certain that the station log  book,  strip charts, etc. are  properly
      dated and signed.                                                               *
                                  2-8

-------
                                  SECTION 3
                          QUALITY ASSURANCE PROGRAM

3.1  INTRODUCTION
     Throughout the  project several quality assurance activities were  per-
formed.   These  included  a  daily zero and span check,  a  two point precision
check  that  was  performed twice  a  week,  two   systems  audits  that  were
performed by the EPA contractor Research Triangle Institute (RTI) during the
project, and a data validation, performed on the  data before it was  reduced.
The following sections will detail  the above mentioned QA activities.

3.2  ZERO AND SPAN CHECKS
     Each day  during the  project  the monitoring sites were  visited  by the
local operator  to  perform  the daily checks.  These  checks  were recorded on
the Daily Check Sheet.
     The zero/span system  utilized a needle flow controller and a system of
"quick connect" valves to  connect  the appropriate cylinder to the analyzer.
The  zero  and   span  values  obtained were  marked on the  strip chart  and
recorded in the operators log book and on the Daily Zero and Span Data Sheet
(Figure 2).  Any  time the  zero or span values were ±10 percent of the cali-
bration zero (including  5% offset) or span  values,  the  local operator con-
tacted the  PEDCo  project manager in Cincinnati  and  then  the instrument was
recalibrated.
     All  zero   and  span  data  has   been  sent  to the  project officer  in  a
separate volume.

                                      3-1

-------An error occurred while trying to OCR this image.

-------An error occurred while trying to OCR this image.

-------
3.3  PRECISION CHECKS



     A two point  precision  check for the NMOC  analyzers  was carried out at



each site twice a week.  These checks were performed after 9 a.m.  EST and on



Mondays and Thursdays.   The precision check data was  recorded  on Precision



Check  Data Forms (supplied  by U.S.  EPA).  These data  are tabulated  in



Appendix B.  The  precision  check data was  used  in  addition to the zero and



spans as an indicator for when recalibration was necessary.





3.4.  SYSTEMS AUDIT



     The NMOC monitoring  network was audited twice during the  study period



by the EPA contractor Research Triangle Institute (RTI).   Table 2 lists the



audit results for the sites.   Only the St.  Vincent site was found to have a



response greater  than 15 percent  of the  audit value.   The instrument was



recalibrated the following day.





3.5  DATA VALIDATION AND AUDIT



     PEDCo's field operator removed,  on a bi-weekly basis, the strip charts



from each of the  four instruments.   The data was  then validated and turned



over to the local  agency for reduction and entry into the State's data base.



     Every effort was made to accomplish this task within a two-week period,



however,  instrument  malfunctions prevented the  field  operator  from keeping



to the bi-weekly schedule.



     Data was invalidated when the  zero, span  or  precision point data were



found to have drifted by more than fifteen percent.
                                      3-4

-------
                        TABLE 2.  NMOC AUDIT SUMMARY
Site/date
Rickoff
School
6/11/81



Rickoff
School
8/14/81



St. Vincents
6/11/81




St. Vincents
8/14/81




University
CAM Site
6/23/81



University
CAM Site
8/11/81



CH4
Audit value,
ppm
0.00
0.98
2.09
3.99
6.07
9.33
0.00
0.99
2.02
3.86
5.84
8.90
0.00
0.97
2.09
3.97
5.96
9.47
0.00
0.96
2.03
3.95
6.07
8.99
0.00
1.00
2.05
3.88
5.98
9.44
0.00
0.96
2.00
3.87
6.01
9.28
Response,
ppm
0.06
1.00
2.12
4.07
6.25
9.41
-0.09
0.89
1.90
3.74
5.73
8.65
0.00
0.95
2.15
4.20
6.35
9.48
0.00
0.94
2.01
3.97
6.06
9.14
0.00
1.00
2.03
3.98
6.06
9.50
0.02
0.98
2.04
4.00
6.13
9.32
Percent
diff.
_.
+2.0
+1.4
+2.0
+2.9
+0.86
—
+10.1
-5.9
-3.1
-1.8
-2.8
--
-1.95
+2.9
+5.8
+6.5
+0.1
--
-2.08
-0.99
+0.51
-0.16
+1.79
--
0.0
-0.98
+2.5
+1.3
+0.6
—
+2.08
+2.0
+3.36
+2.0
+0.43
NMOC
Audit value,
ppm
0.00
0.33
0.65
1.27
1.94
3.04
0.00
0.32
0.65
1.23
1.86
2.84
0.00
0.31
0.66
1.28
1.95
3.06
0.00
0.31
0.65
1.26
1.94
2.88
0.00
0.32
0.65
1.24
1.94
3.05
0.00
0.31
0.65
1.28
1.99
2.96
Response,
ppm
0.03
0.34
0.66
1.31
1.98
3.11
-0.01
0.30
0.64
1.24
1.88
2.85
0.00
0.35
0.77
1.43
2.15
Off scale
-0.02
0.30
0.64
1.26
1.92
2.88
0.00
0.32
0.66
1.25
1.96
3.07
0.01
0.33
0.68
1.31
2.00
2.92
Percent
diff.
__
+1.5
+1.7
+3.1
+2.1
+2.3
--
-6.3
-1.5
+0.8
+1.1
+0.4
..
+12.2
+16.7
+11.7
+10.3
—
--
-3.2
-1.5
0.00
-0.91
0.00
--
0.0
+1.54
+0.8
+1.0
+0.7
—
+6.06
+4.6
+2.3
+0.5
+0.43
(continued)
                                     3-5

-------
TABLE 2 (continued)
Site/date
Norwood Sit«
6/23/81




Norwood Sit«
8/12/81




CH4
Audit value,
ppm
0.00
0.99
2.04
4.45
6.79
9.04
0.00
0.93
2.01
3.92
5.92
8.85
Response,
ppm
0.00
1.02
2.08
4.49
6.74
9.18
0.12
0.87
1.90
3.82
5.87
8.64
Percent
diff.
—
+3.0
+1.96
+0.9
-0.7
+1.5
—
-6.45
-5.47
-2.55
-0.84
-2.37
NMOC
Audit value,
ppm
0.00
0.32
0.65
1.27
1.93
2.89
0.00
0.30
0.64
1.25
1.89
2.17
Response,
ppm
0.00
0.00
0.66
1.29
1.95
2.91
0.01
0.32
0.66
1.30
1.96
2.23
Percent
diff.
__
—
+1.5
+1.6
+1.0
+0.7
—
+6.67
+3.13
+4.00
+3.70
+2.77
                                     3-6

-------
                                  SECTION 4



                           SITE OPERATION SUMMARY





4.1  RICKOFF SCHOOL SITE



     The NMOC analyzer at this site was initially calibrated and put on line



on June 11,  1981.   The instrument operated throughout the study without any



problems with the  exception  of a one week period  in late June.   The flames



went out  on June  17  and were not able  to  be relit until June  19.   It was



determined  that this  problem was due  to a  pump  malfunction which  caused



fluctuating pressures.



     On June 22  there was a power failure which caused the loss of approxi-



mately  one  day's worth  of  data.  After  these problems  were  corrected the



analyzer operated without incident.





4.2  ST. VINCENT'S SITE



     The St. Vincent's  site  was initially calibrated on  June  4, 1981.  The



only problem  experienced at  this  site was  the failure  of  the temperature



control board for  the oxidizer.   A new control board was ordered on July 1,



1981 and installed on July 14, 1981.   After this was corrected there were no



other problems at this site.





4.3  UNIVERSITY CAM SITE



     The University  CAM site  was  initially  calibrated  and put  on line on



June 1, 1981.   This  site operated without any major problem for the entire



study period.  The only minor problem was with the NMOC strip chart on June



2, 1981.






                                      4-1

-------

4.4  NORWOOD SITE



     The Norwood  site was  initially  calibrated  and put on line on  June  8,



1981.  The  site operated  without  any major problems until August 4,  1981.



At  this  time  the  instrument  drifted  out  of  balance.   Due  to  scheduling



conflicts the analyzer was not recalibrated until August 10, 1981.



     On  September 4,  1981 several  hours  of  data were  lost when  the  CH4



channel strip chart was allowed to run out.



     The air  conditioner  at the  site went out on  September  25,  1981.   The



analyzer was  shut  down until  September 26 to prevent damage to the analyzer



due to overheating.



     Other  than these  few problems,  the  analyzer operated  throughout  the



study period.
                                      4-2

-------
        APPENDIX A

 LOCAL OPERATORS STANDARD
OPERATING PROCEDURES (SOP)

-------
                             ZERO AND SPAN CHECK
                     STANDARD OPERATING PROCEDURE (SOP)
                           MSA-11-2 NMOC ANALYZER
     After completing  the daily checklist,  begin the  zero  and span  check

according to the following stepwise procedure.

     1.    Check moisture  trap No.  1  on  the  side of the  instrument.   If it
          contains more than  1"  of water,  open valve SLOWLY  and drain to a
          level of 1/4".   Do not drain completely.

     2.    Check moisture trap No. 2, also on  the side of the  instrument.   If
          it contains  more than 1/2" of water,  open  the toggle  valve  and
          drain.

     3.    Check both  pressure gauge  (No. 1  and No.  2).  No.  1  should read
          approximately 40 psig.   No.  2 should read 10 psig less than No. 1,
          adjust No.  2 if it does not.

     4.    Disconnect the  "pump  inlet" line  from the sample manifold to  the
          vacuum  side  of the pump.   Then  connect the  "T-line"  to the same
          fitting.   CAUTION:   Do  not over-tighten  the  connection, finger
          tight and then a 1/4 turn with the  wrench should suffice.

     5.    Connect the zero stem to the "quick connect" side of the "T-line".
          Check for  excess  flow through the  open end of the "T-line".  Flow
          through the  "T-line"  may be increased by  turning the  knob on the
          face of the regulator clockwise.   Decrease the flow by turning the
          knob counter-clockwise.

     6.    Allow  sufficient time  (approximately  5  to  10 minutes)  for  the
          analyzer to  stabilize.   Record the zero value on the daily check-
          list, the  daily zero/span  data  sheet  and plot the  value on the
          zero/span  chart.   Note  - the  pressure on the  span  and  zero tank
          should be 60 lb/in.2

     7.    Disconnect  the  zero  stem  and  connect  the  SPAN stem  to  the
          "T-line".   Check for  excess  flow through  the open  end  of  the
          "T-line".   Flow through  the "T-line" may  be  increased by turning
          the  knob  on the  face of the  regulator clockwise.   Decrease  the
          flow by turning the knob counter-clockwise.

     8.    Allow  sufficient time  (approximately  5  to  10 minutes)  for  the
          analyzer to  stabilize.   Record the span value on the daily check-
          list, the  daily zero/span  data  sheet  and plot the  value on the
          zero/span chart.

                                      A-2

-------
A-3

-------
 9.    If either value  for  both Channel 1 and  Channel  2 are off by more
      than the 10 percent indicated on the control  chart, call  the PEDCo
      Project Manager, Ed Mull in (513) 782-4700 immediately.

10.    Disconnect the  SPAN  stem from the "T-line".   If a precision check
      is to be  performed  at this time, continue with the SOP for preci-
      sion checks, if not,  then proceed to step 11.

11.    Disconnect the "T-line" from the vacuum side of the pump.   Connect
      the pump  inlet  line  to the pump  fitting.  CAUTION:   Do  not over-
      tighten the connection,  finger  tight  and a  1/4  turn  with  the
      wrench should suffice.

12.    Label both  zero and  span points on the  strip  charts,  record the
      time and date and sign both strip charts.
                                  A-4

-------
                               PRECISION CHECK
                        STANDARD OPERATING PROCEDURE
     The following  procedure for  precision  checks  is to follow  step  10 of

the ZERO and SPAN check SOP.

     1.    Connect the precision check gas cylinder No. 1 (Precis.  Ck.  No. 1)
          stem to the  quick  connect side of the "T-line".   Check for excess
          flow  through the  open  end  of  the  "T-line".   Flow through  the
          "T-line" may be  increased by turning the knob on  the face of the
          regulator  clockwise.    Decrease  the  flow   by  turning  the  knob
          counter-clockwise.   Allow sufficient time  (approximately 5  to 10
          minutes)  for the  analyzer  to stabilize.   Record  the  precision
          check point  No.  1 valve on  the form and label the  strip  charts.
          Note  -  the  pressure  for  the  precision  span  tanks  should  be 60
          lb/in.2

     2.    Disconnect the Precis.  Ck.  No. 1 stem and connect the Precis.  Ck.
          No.  2  stem to the "T-line".   Check  for  excess  flow through the
          open  end  of the  "T-line".   Flow  through the  "T-line"  may be
          increased by turning  the knob  on the face of the regulator clock-
          wise.    Decrease  the   flow  by turning the knob counter-clockwise.
          Allow  sufficient  time  (approximately  5  to  10  minutes)  for  the
          analyzer  to  stabilize.   Record the precision  check point  No.  2
          valve on the form and label the strip charts.

     3.    Disconnect the Precis. Ck. No.  2 stem from the "T-line".

     4.    Disconnect the "T-line" from the vacuum side of the pump.  Connect
          the "pump inlet"  line  to  the  pump fitting.   CAUTION:  Do  not
          overtighten the connection, finger tighten and a 1/4 turn with the
          wrench should suffice.

     5.    Label  both precision  points  on the strip charts,  record the time
          and date,  and  sign both strip  charts.  NOTE:   If a zero and span
          check was  also performed,  label these points on the strip charts.
                                      A-5

-------
                              RELIGHTING BURNER
                        STANDARD OPERATING PROCEDURE


     If the flame  out  indicator is ON (the light will be on), the burner(s)

will need to be relit.   The following stepwise procedure should be completed

before calling the  PEDCo  Project Manager, Ed Mullin (513) 782-4700.   If the

burner does not relight easily, call the PEDCo Project Manager.

     1.   Adjust the AIR  regulator (on the face of  the instrument)  to 5 to
          10 psig.

     2.   Press the  IGNITION button  (also the  flame  out indicator)  for a
          count of  three.   If the burner has been out a while, it will take
          several  attemps to successfully relight the burner.

     3.   If the burner does not ignite on the first try, repeat 3 or 4 more
          times.    NOTE:   It will  take several seconds for  the  light to go
          out after the burner ignites.

     4.   Adjust the AIR regulator to 20 psig.

     5.   Record on all  the strip charts which  channel  was  relit and when.
                                      A-6

-------
       APPENDIX B



PRECISION CHECK PROGRAM

-------
Site:   A.  J.  Rickoff School
Week of
6/29/81
7/6/81
7/13/81
7/20/81
7/27/81
8/3/81
8/10/81
8/17/81
8/24/81
8/31/81
Precision
check No.
1
1
2
3
1
2
1
2*
1
2
I
2
1
2
3*
4
1
2
1
2
1
2
Cylinder No. LL 3710
NMOC,
(ppm)
0.54
0.64
0.54
0.54
0.62
0.53
0.61
0.50
0.52
0.47
0.49
0.50
0.46
0.44
0.58
0.52
0.44
0.42
0.41
0.40
0.42
0.55
CH4,
(ppm)
1.52
1.66
1.64
1.64
1.56
1.57
1.45
1.61
1.55
1.73
1.61
1.75
1.61
1.61
1.50
1.41
1.46
1.52
1.56
1.61
1.41
1.63
       Precision gas cylinder was empty.
                         B-2

-------
Site:   St.  Vincent's
Week of
6/15/81
6/22/81
6/29/81
7/6/81
7/13/81
7/20/81
7/27/81
8/3/81
8/10/81
8/17/81
8/24/81
9/1/81
9/14/81
9/21/81

Precision
check No.
1
1
2
1
2
3*
1
2
1
2
1
2
3*
1
2
1
2
1
2
1
2
1
2
3
1*
1
2
1
2
1
Cylinder No. LL 3709
NMOC,
(ppm)
0.45
0.46
0.42
0.41
0.44
0.51
0.50
0.49
0.47
0.48
0.46
0.47
0.52
0.52
0.50
0.51
0.52
0.51
0.50
0.50
0.50
0.50
0.47
0.48
0.53
0.49
0.48
0.50
0.48
1 	 0.45
CH4,
(ppm)
1.59
1.61
1.63
1.60
1.53
1.64
1.70
1.68
1.71
1.76
1.78
1.73
1.62
1.68
1.68
1.63
1.72
1.71
1.56
1.77
1.77
1.74
1.73
1.66
1.67
1.70
1.63
1.68
1.76
1.73
          *New calibration relationship.
                         B-3

-------
Site:   University CAM
Week of
6/22/81
6/29/81
7/6/81
7/13/81
7/20/81
7/27/81
8/3/81
8/10/81
8/17/81
8/24/81
8/31/81
9/7/81
9/14/81
9/21/81
9/30/81
Precision
check No.
1
2
1
2
3
1
2*
3
1
2
3
1
2
3*
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
2
3
1
1
1
Cylinder No. LL 4851
NMOC,
(ppm)
0.52
0.50
0.52
0.53
0.54
0.53
0.48
0.45
0.47
0.46
0.47
0.46
0.47
0.50
0.47
0.48
0.49
0.50
0.48
0.50
0.49
0.50
0.49
0.47
0.52
0.51
0.55
0.50
0.50
0.50
0.50
0.49
0.51
0.54
0.53
0.52
0.50
0.46
CH4f
(ppm)
1.64
1.63
1.63
1.59
1.63
1.63
1.64
1.59
1.63
1.64
1.60
1.59
1.59
1.64
1.68
1.64
1.64
1.64
1.63
1.63
1.64
1.63
1.64
1.63
1.64
1.64
1.65
1.63
1.63
1.63
1.63
1.63
1.55
1.66
1.65
1.63
1.63
1.63
          *New calibration relationship.
                         B-4

-------
Site:   Norwood
Week of
6/22/81

6/29/81



7/6/81


7/13/81


7/20/81



7/27/81


8/3/81


8/10/81


8/17/81


8/24/81


8/31/81


9/6/81


9/14/81
9/21/81
Precision
check No.
1
2
1
2
3
4
1
2*
3
1
2
3
1
2*
3
4
1
2
3
1
2
3
1
2*
3
1
2
3
1
2
3
1
2
3
1
2
3
1
1
Cylinder No. LL 3800
NMOC,
(ppm)
0.55
0.55
0.56
0.55
0.52
0.55
0.55
0.55
0.55
0.54
0.53
0.47
0.54
0.53
0.54
0.55
0.57
0.53
0.54
0.55
0.35
—
--
0.57
0.53
0.51
0.50
0.52
0.55
0.52
0.54
0.56
0.53
0.53
0.53
0.54
0.53
0.55
0.54
CH4,
(ppm)
1.63
1.64
1.59
1.53
1.50
1.54
1.56
1.62
1.62
1.67
1.55
1.60
1.56
1.62
1.62
1.60
1.60
1.59
1.59
1.59
1.66
1.57
1.59
1.45
1.49
1.49
1.51
1.47
1.44
1.46
1.50
1.50
1.50
1.55
1.50
1.47
1.65
1.50
1.49
          *New calibration relationship.



                         B-5

-------
TECHNICAL REPORT DATA
(Please read Instructions on the reverse before completing/
iTREPORT NO.
FPA-905/4-82-002
2.
4. TITLE AND SUBTITLE
Non-Methane Organic Compound Continuous Mon
in Cleveland and Cincinati: 1981 Ozone Mon
itoring
i tor ing
Study
7- AUTHOR(S)
i
ki.
E. W. Mullin, Jr.



jf PERFORMING ORGANIZATION NAME AND ADDRESS
PEDCo Environmental, Inc.
1
1499 Chester Road


Cincinnati, Ohio 45246
12. SPONSORING AGENCY NAME AND ADDRESS
U.S. Environmental Protection Aaencv
Region V


536 Clark Street
Clhicaqo, Illinois 60605
15.


16.
SUPPLEMENTARY NOTES





3. RECIPIENT'S ACCESSION NO.

5 REPORT DATE
August. 1QR?



6. PERFORMING ORGANIZATION CODE


8. PERFORMING ORGANIZATION REPORT NO.
PN 3525-22

10. PROGRAM ELEMENT NO.


11. CONTRACT/GRANT NO.
68-02-3512
13. TYPE Of REPORT AND PERIOD COVERED
Final
14. SPONSORING AGENCY CODE








EPA Project Officer: Stephen K. Goranson (312) 886-6229
ABSTRACT


This program was proposed to obtain Non-Methane Organic


Compound (NMOC) data for
djays of high ozone concentration during the summer of 1981. Data were especially
nleeded in the vicinity of Cleveland and Cincinnati to provide the needed input
into the productive model that will be used to demonstrate attainment
of the
ambient ozone standard.

17.
a.





KEY WORDS AND DOCUMENT ANALYSIS
DESCRIPTORS

Air pollution organic compounds
18.

i
DISTRIBUTION STATEMENT
Unlimited





b. IDENTIFIERS/OPEN ENDED TERMS
Monitoring
Non-Methane
Compounds
Cleveland
Cincinnati
Organic
19. SECURITY CLASS (This Report)
Unclassified
20. SECURITY CLASS (This page/
Unclassified
C. COSATI Field/Group
13B
07C
21. NO. OF PAGES
32
22. PRICE

»A Form 2220-1

-------