United States
                   Environmental Protection
                   Agency
Atmospheric Research and
Exposure Assessment Laboratory
Research Triangle Park NC 27711
                   Research and Development
 EPA/600/S3-89/070  Feb. 1990
&EPA         Project  Summary
                  Analysis of Ambient Polar
                  Volatile Organic Compounds
                  Using Chemical lonization-lon
                  Trap  Detector
                  Sydney M. Gordon and Michael Miller
                    The current approach to measuring
                  trace levels  of  volatile  organic
                  compounds (VOCs) in ambient air
                  requires cryogenic trapping of the
                  analytes, followed by thermal desorp-
                  tion and low-temperature refocussing
                  onto a  column  for  analysis by
                  capillary gas chromatography/mass
                  spectrometry (GC/MS). This approach
                  has been  successfully applied to
                  nonpolar VOCs, but its use for more
                  polar species has been complicated
                  by the problems associated with the
                  ambient water vapor collected  with
                  the VOCs.
                    A promising technique for meas-
                  uring polar VOCs is chemical ion-
                  ization GC/MS (CI-GC/MS)  in the
                  quadrupole  ion trap. This  approach
                  allows whole air samples to be taken
                  since the water present in the air is
                  used as the Cl reagent gas. Water Cl
                  leads to appreciable intensities for
                  the proton transfer agent H3O*,
                  which produces intense pseudomo-
                  lecular  ions and class-specific  frag-
                  mentation patterns for various low
                  molecular weight polar compounds.
                  Using water Cl enhances sensitivity
                  by a factor of about 5, as is apparent
                  from a comparison of Cl and electron
                  impact  intensities  obtained with the
                  Finnigan MAT 800 Ion Trap Detector™
                  (ITD) in full scan mode.
                    Standard mixtures of polar species
                  at low concentrations in  humidified
                  zero  air were analyzed without a
                  membrane  dryer, using a cryogenic
                  trap and GC/CI-ITD with the ion trap
                  detector in  the full scan mode. Water
                  appears to  be an effective Cl reagent
gas in the ion trap, and the GC/CI-ITD
system can exploit the water vapor
normally present in ambient  air to
circumvent the problems usually en-
countered  when  analyzing humid
ambient air for these compounds.
The  water  vapor  present  in  the
samples does, however, have a dele-
terious effect on the quality of  the
chromatography achieved.  Future
work will focus  on optimizing  the
analytical technique and evaluating
its suitability for  various polar com-
pound classes.
  This Project Summary was  devel-
oped by EPA's Atmospheric Research
Exposure Assessment Laboratory,  Re-
search Triangle Park, NC, to announce
key findings of the research project
that is fully documented in a separate
report of the same title (see Project
Report ordering information at back).

Introduction
  Increasing public concern over polar
volatile organic compounds  (VOCs) of
low molecular weight has spurred  the
current  interest in developing  better
measurement techniques for  these
compounds.  Some  polar VOCs are often
the subject of nuisance complaints from
the public (e.g., alkylthiols, amines, alde-
hydes, etc.),  while others are cited as
potentially toxic compounds from indus-
trial emissions (e.g., ethylene  oxide,
propylene oxide, acrolem, etc.). Analysis
of  trace levels of polar VOCs  in  air,
however, poses many problems,  and at
present, the nature and  distribution of
these compounds in the atmosphere
cannot be adequately characterized.

-------
  By  contrast,  the  method for  deter-
mining a large number of nonpolar VOCs
in  air  is well established. Stainless steel
polished sampling canisters  are  used
with a combination of cryogenic trapping
and high resolution gas chromatography
(GC) coupled to a suitable detector, such
as flame ionization, electron capture,  or
mass  spectrometry (GC/MS).  Cryogenic
trapping preconcentrates the analytes by
first passing the gaseous sample through
a cold tube  in  which  the organic  com-
ponents condense on  the cold surface;
the condensate is  then desorbed  and
analyzed.
  In general,  attempts  to sample  polar
VOCs  at trace  levels  have been inef-
fective due to their  chemical  reactivity,
affinity for surfaces,  and  tendency  to
undergo polymerization. The  analytical
problem has  been exacerbated,  espe-
cially  for GC/MS measurements, by co-
collected  water  vapor,  which  occurs  at
concentrations that are several orders  of
magnitude greater than  that of the
analytes.
  The moisture  problem has been over-
come for many  nonpolar compounds by
incorporating a  drying  step before GC
analysis. The air sample passes through
a  Nafion  membrane dryer  tube before
being  cryogenically concentrated;  water
vapor in the air stream permeates the
dryer  walls  while  most  of the  air
pollutants  of interest  pass through the
tube. This  technique is limited, however,
to  relatively  nonpolar  compounds  since
small  polar molecules also permeate the
walls of the drying tube.
  A promising  technique for measuring
volatile polar  compounds in  whole  air
samples is chemical  ionization  GC/MS
(GC/CI-MS) in  the quadrupole ion trap,
using  the water  present in the air as the
chemical ionization reagent gas. Water  Cl
has been  shown to  produce intense
pseudomolecular ions  and class-specific
fragmentation patterns for several polar
compound classes  in a conventional
quadrupole GC/MS system.  To date, the
chemical  ionization  capabilities of the
Finnigan MAT Ion  Trap Detector have
been investigated in detail for the reagent
gases ammonia, methane, and  isobutane.
A  recent study has shown that chemical
ionization  in the trap  can  improve the
lower  detection limit  for  some  com-
pounds by as much as  a factor of 10 over
that for electron impact (El)  ionization.
The relatively low cost and exceptional
sensitivity  of the  ITD  make it an ideal
candidate  detector for the  analysis  of
polar VOCs of environmental interest.
  The  objectives of this study were  to
evaluate  the conditions  under  which
water chemical ionization can be used for
the analysis of selected polar VOCs in
whole air  samples,  using  GC/CI-ITD  in
the full  cyclic scan  mode.  The  study
consisted  of generating  and  comparing
electron  impact  and water Cl  mass
spectra  for several  representative polar
compounds, and evaluating the effects of
humidified air and dry air on  the CI-ITD
analysis of  these compounds.

Procedures

Sampling System
  The system used  for  sampling and
analyzing volatile  organic compounds in
zero-grade  air streams  consisted of  a
semi-automated cryogenic trapping and
desorption  unit (Nutech  Model 3538-02)
and  a  Fmnigan  MAT  800 Ion Trap
Detector. Normally,  whole  air samples
collected in stainless steel canisters  are
preconcentrated  by  passing  them
through  the cryogenic trapping system
prior to  analysis. To  investigate  the
effects of ambient water vapor on  the
analysis of  polar VOCs, a special sample
delivery system  was constructed. This
system,  which  was  used  in lieu  of
canisters for whole-air samples, allowed
us to generate sample mixtures of polar
compounds in  dry zero air or humidified
zero air. Samples were injected into a dry
zero-grade  air  stream through a septum.
Humidified  samples were prepared by
first  passing  the air through  a flask
partially  filled  with  water,  while dry
samples  were generated by  bypassing
the humidifier. Experiments  were  also
carried out, using  a Nafion membrane
dryer tube, to demonstrate the effects of
removing water from  a  humidified gas
stream on the analysis of polar VOCs.

Instrumentation
  A Varian  3400 GC, equipped with a 60
m x 0.32  mm id fused  silica capillary
column  (J&W DB-1701)  with  a film
thickness of 0.2 jim, was attached directly
via a heated  transfer line  to  a Finnigan
MAT 800 Ion  Trap  Detector  (ITD). The
column  was temperature programmed
from -10°C at 10°C/min to 250°C. To
avoid problems with humidified samples
that caused ice to form in the column and
block it, a 1m x 0.53 i.d.  fused silica pre-
column  was connected to the analytical
column inlet.
  The Finnigan MAT 800  ITD is a fully
integrated benchtop mass spectrometer
with  high sensitivity in both the electron
impact (El) and  the  chemical ionization
(Cl) mode.  It is also  capable of full cyclic
scanning and selected ion monitoring. AJ
opposed to  Cl  in  a conventional mass
spectrometer ion source, Cl in the  ITC
requires only 10-6 to 10'4 torr of reagem
gas.  This low pressure is  sufficient tc
convert  sample molecules  to  ions with
high efficiency  because of  the  long
reaction times (milliseconds) in the trap
The ion trap operates in a pulsed mode
a special sequence  of  rf voltages  anc
time intervals are  used to select  the
reagent ions, then react them with  the
sample to form analyte  ions before the rl
voltage is scanned for mass analysis.

EI/CI Mass Spectra Study
  For the water Cl experiments, an exter-
nally mounted water source was used tc
maintain  a constant reagent gas pressure
in the ion trap. The water was introduced
into the ion  trap from a vial attached tc
the  reagent gas inlet port. Several repre-
sentative polar compounds were injected
into the instrument to obtain their El  and
water Cl mass spectra.

Trapping and Recovery Studies
  Cryogenic trapping  and  recovery
studies of several polar VOCs of interest
(and some  nonpolar  compounds,  for
comparison) were carried  out  using the
sample  delivery  system  and the cryoi
genie trapping unit described  above.
Using a gas-tight syringe, a fixed volume
of  a standard mixture  was injected  into
the  humidified zero-grade air stream in
the  sample  delivery  system. The VOCs
were concentrated in the cryogenically
cooled  trap. To  desorb the trapped
sample, the  temperature of the trap  was
raised rapidly and the contents of the  trap
were flushed onto the capillary column
for GC-ITD analysis. Several experiments
were carried out  under dry  conditions
(i.e., no  water  in  the  humidifier),  and
some  runs were  performed using
humidified air with the Perma-Pure dryer
between  the sample delivery system  and
the trapping  unit.

Results and Discussion

Comparison of El and Water Cl
Mass Spectra
  Some  of  the water  Cl  and El mass
spectra obtained for selected polar VOCs
using fixed operating conditions in the ion
trap  are summarized in Table  1.  The
reduction in molecular  ion fragmentation
achieved by using water Cl instead of El
is  striking. All of the water Cl spectra,
except for  amyl  acetate,  have  thd
(M + H)*  pseudomolecular ion as one of

-------
  fable 1.    Ion  Trap  Cl  and El Mass  Spectra  of
            Selected Polar VOCS
                              % Pel. Abundance
Compound
3-Octanone
(MW 128)





Benzophenone
(MW 182)






Acroleirt
(MW 56)

Amyl Acetate
(MW 130)







Heptanal
(MW114)










m/z
129
99
73
72
71
57
55
184
183
182
181
105
77
51
50
57
55
56
71
70
61
55
43
42
41
40
39
115
97
81
71
70
69
68
67
57
55
53
45
Water Cl
100




10

13
100


47



100


15

100

35




23
100


25
17


15
46


El

35
16
34
65
100
16


41
10
100
86
67
29

100
32

11
15
14
100
20
20
12
15

97
36
18
62
17
20
18
50
100
10
22
 the major  ions.  Heptanal  is  the only
 compound  listed  in which the (M + H)*
 ion does not occur as the base peak.
   Benzophenone  and  naphthalene were
 used to compare the sensitivity of  the
 ITD in  the  full cyclic  scan  mode under
 both El and  water Cl conditions.  For
 naphthalene, which undergoes  very little
 fragmentation by  either El or  water Cl,
 the estimated limit of detection was about
 0.05 ng. For benzophenone,  which frag-
 ments to a  much larger extent by El, the
 smallest amount of sample  required to
 give good quality  full scan mass  spectra
 (mass range m/z 45-300) was about 0.03
 ng by water Cl  and 0.15  ng  by El  This
^enhancement in  sensitivity using  Cl
 (irises because the ion current  is  usually
"concentrated in  only a few fragments in
Cl,  as opposed to the large  number  of
fragments involved  in  the  El  case.
Furthermore, the increased reaction time
in the CI-ITD mode results in a greater
accumulation  of  sample ions for  detec-
tion.
Water CI-ITD Analysis of
Humidified and Dry Air Samples
  A standard mixture containing selected
polar  and nonpolar  VOCs  was analyzed
by water Cl in the  ion trap, to evaluate
the effects of humidified  zero-grade air,
dry zero-grade air, and Nafion dried air
on  simulated  whole-air  samples. The
chromatographic  behavior of the com-
pounds listed was generally quite poor.
This was largely  due  to the fact that the
water vapor had a significant effect upon
the chromatography  of the compounds,
and that neither the GC column nor the
operating conditions  were optimized  for
this work. Nevertheless, the experiments
carried out were able to demonstrate the
value of the water Cl technique for the
analysis of polar VOCs.
  Despite the  effects of the water vapor
on the  chromatography,  relatively little
retention time  variability was observed in
these experiments.  Table 2 shows the
average  retention times obtained for the
test  mixture  that was analyzed  under
humidified, dry  and  Perma-Pure dried
conditions. The increased fluctuations
noted with shorter retention times in the
table has been observed before for non-
polar compounds,  and  occurs  when
cryogenically  trapped compounds  in-
cluding  water are  transferred  from the
trap to a capillary column held  initially at
a reduced temperature. Despite the use
of a precolumn and a slow initial temper-
ature  program, the  column is probably
blocked  temporarily  by  ice  until  the
column  temperature rises  above  the
freezing  point of the water. Since  this
condition is difficult to control due to the
erratic character of the blocking process,
the retention  time fluctuations for the
early  eluting  compounds are not unex-
pected
  The effect of the Perma-Pure dryer on
some of the  compounds  in  Table 2  is
striking. All of  the polar compounds, with
three  exceptions, are removed from the
gas stream along with the water  vapor.
The exceptions are  acetone  (very small
peak),  acrylonitrile, and 1-methyl-2-pro-
panethiol.  Both  acetone  and  acryloni-
trileare soluble in water. Thus,  the small
amount of acetone  detected is most
likely  due  to  the fact that acetone was
present at a very high level in the original
mixture.  The disappearance of limonene
from this sample is equally  surprising,
since  it is  insoluble in water and  should
have passed  through  the  dryer into  the
GC column, in much the  same way  as
styrene  or m-dichloro-  benzene.  The
reasons for this  aberrant behavior in  the
case of acrylonitrile and limonene are not
understood at this time.
Conclusions and
Recommendations
  The  laboratory  tests that  have  been
conducted to evaluate the effects of
humid  air matrixes on the  analysis of
polar  VOCs  using water Cl  with the
Finnigan MAT Ion Trap Detector, indicate
the following:

-------
Table 2.   Retention Times  for  Several  Polar  and  Nonpolar Compounds
          Cryogenically Collected and Analyzed by Water CI-ITD
                                     Retention Time, minutes
Compound
Acetone
Acetonitrile
Acrylonitrile
Tetrahydrofuran
Ethyl aceiate
2-Butanone
1 -Methyl-2-propanethiol
Acrolein
Styrene
Amyl acetate
LJmonene
m-Dichlorobenzene
Nitrobenzene
Humidified*
8.66
10.11
10.73
12.89
12.97
13.20
16.68
18.39
23.49
24.12
25.77
26.21
29.43
Dry
8.22
968
10.55
12.68
12.83
13.00
16.57
18.15
23.45
24.10
25.75
26.18
29.38
Perma-Pure
Dried
9.05
ND*>
10.83
NO
ND
ND
16.70
ND
23.22
ND
ND
26.25
29.50
"Average of two runs
bND = not detected
                                              •  Water  appears to be  an effective  Cl
                                                 reagent gas in the  ion trap.  Its proton
                                                 affinity is low enough to allow it to react
                                                 with a wide range of polar VOCs.
                                              •  The ITD permits the use of the very long
                                                 reaction times necessary to detect low
                                                 sample  concentrations  with  high
                                                 sensitivity.
                                              •  The GC/CI-ITD system can  exploit the
                                                 water  vapor  present  in  the air,  thus
                                                 circumventing the  problems  normally
                                                 encountered when analyzing humid air.
                                                 However, the presence  of water vapor in
                                                 the air has a deleterious  effect  on the
                                                 quality of the chromatography achieved.
                                                 The exceptional sensitivity of the  Ion
                                              Trap Detector, together with water chemical
                                              ionization,  may lead  to  the use  of
                                              significantly smaller  whole air samples to
                                              further reduce the deleterious effects of
                                              humidified air on the trace-level analysis of
                                              polar VOCs.
 Sydney M.  Gordon and Michael Miller are with IIT Research Institute,  Chicago, IL
   60616-3799.
 Joachim Pleil is the EPA Project Officer (see below).
 The  complete report, entitled  "Analysis  of Ambient Polar Volatile  Organic
   Compounds  Using Chemical lonization-lon Trap Detector," (Order No. PB 90-106
   451 /AS; Cost: $15.00, subject to change) will be available only from:
         National Technical Information Service
         5285 Port Royal Road
         Springfield, VA 22161
         Telephone: 703-487-4650
 The EPA Project Officer can be contacted at:
         Atmospheric Research and Exposure Assessment Laboratory
         U.S. Environmental Protection Agency
         Research Triangle Park, NC 27711
 United States
 Environmental Protection
 Agency
Center for Environmental Research
Information
Cincinnati OH 45268
                                                                                             X  '6NALTV
                                                                                              US.OFFICIAL MAIL
U.S.POSTAG1

= 0  .2  5  H
  Official Business
  Penalty for Private Use $300

  EPA/600/S3-89/070

-------