United States
                   Environmental Protection
                   Agency
Environmental Monitoring and
Support Laboratory
Cincinnati OH 45268
                    Research and Development
EPA/600/S4-88/030 Sept. 1988
x°/EPA         Project  Summary
                   Capillary  Column  GC-MS
                   Determination  of  77  Purgeable
                   Organic Compounds  in  Two
                   Simulated   Liquid  Wastes

                   M. F. Yancey, R.A. Kornfeld, and J. S. Warner
                     The suitability of purge-trap-desorb
                   (PTD) procedures for determination of 84
                   volatile  organic  compounds with
                   capillary column gas chromatography
                   (GC) and mass spectrometry (MS) was
                   evaluated. After collecting GC-MS data
                   not previously available  for  some
                   analytes, 7 of the 84 compounds were
                   eliminated from further consideration
                   because of poor purging efficiency or
                   analyte stability problems.
                     For each of the remaining 77 com-
                   pounds, the linear concentration range
                   and detection limit were determined with
                   data obtained by PTD GC-MS analysis of
                   spiked reagent water. A relative standard
                   deviation (RSD) of <25% for the average
                   response factor (RF) was chosen as the
                   acceptance criterion for determining the
                   linear range. This criterion was met over
                   a concentration range of at least two
                   orders of magnitude for  56 of the 77
                   analytes, 1.5 orders of magnitude for 12
                   analytes, and  1 order of magnitude for
                   6 analytes. The criterion was not met for
                   acetone, trichlorofluoromethane, and
                   2-chloro-1,3-butadiene.
                     Method performance was assessed by
                   analyzing eight replicate aliquots of each
                   of two simulated liquid waste samples (a
                   municipal sewage sludge leachate and
                   reagent water containing fulvic acid)
                   containing analytes spiked at two con-
                   centrations. For > 80% of the analytes,
                   bias of measured concentrations was
                   <30%. For most  other  analytes  ac-
                   curacy was > +30%. The observed high
                   positive bias was attributed to enhanced
                   sensitivity caused  by high concentra-
                   tions of ions in the MS source. Calibra-
tion data showed that short term (daily)
and long term (two weeks) precision was
very good.
  This Project Summary was developed
by EPA's Environmental Monitoring and
Support Laboratory, Cincinnati, OH, to
announce key findings of the research
project  that is  fully documented in a
separate report of the same  title (see
Project  Report ordering information at
back).
Introduction

  The Resource Conservation  and
Recovery Act (RCRA) specifies over 200
toxic organic compounds (in Appendix IX
to 40 CFR, Parts 264 and 270) to be used
to screen for suspected ground water con-
tamination at land-based hazardous waste
treatment, storage, and disposal facilities
(Federal Register 52, July 9, 1987, pp.
25942-25953). Analytical methods for most
of these analytes are published in SW-846,
"Test Methods for Evaluating Solid Wastes,
Physical/Chemical Methods," Third  Ed.,
November, 1986).
  The SW-846 method recommended for
determining volatile, relatively water insolu-
ble, organic analytes is Method 8240, which
involves purge-trap-desorb (PTD) analyte
extraction followed by packed column GC
separation and MS detection and measure-
ment. Advances in GC column technology
now permit determination of a wider range
of compounds in a shorter time with greater
sensitivity using a fused  silica or glass
capillary column. In this study, a 0.75 mm
i.d. glass capillary column was used to

-------
evaluate Method 8240 procedures for 84
analytes. The compounds considered for
inclusion in this study include all USEPA
Method 524.2 analytes and all compounds
on the Appendix IX list (Federal Register,
57, 26639, July 24, 1986) that might be
amenable to determination  by  room
temperature PTD extraction followed by
GC-MS analysis using a 0.75 mm i.d. glass
capillary column.

Experimental

PTD-GC-MS Analyses

  Analyses were performed with Method
8240 procedures using a Tekmar Model
LSC-2  PTD system, a Carlo Erba Model
4160 GC, a Finnigan Model 3200 MS fitted
with a glass jet separator, and an Incos data
system with Revision 5.5 software. The PTD
system was fitted with a 5-mL fritted glass
purge tube and a  305 mm x 4 mm i.d.
stainless steel trap containing 10 mm of 3%
SP-2100 on Supelcoport, 77 mm of Tenax,
77 mm of silica gel, and 77 mm of coconut
charcoal. The system was operated with a
helium purge for 11 min at 26 mL/min at
room temperature (23-25°C), desorption for
4 min at 15 mL/min at 180°C, and a trap
bake for 7 min at 26 mL/min at 180°C. The
GC was fitted with  a 60 m x 0.75 mm i.d.
Supelco VOCOL column coated with a 1.5
urn film and operated with helium carrier
gas  flow of 15  mL/min. The  column
temperature was maintained at 10°C dur-
ing the desorb cycle, programmed to 200°C
at 10°C/min at the end of the desorb cycle,
and maintained at 200°C for 10 min. The
MS was tuned daily to meet bromofluoro-
benzene  (BFB)  criteria daily  and was
operated with a scan time of 1 sec over a
mass range of 35-325 amu. The emission
current was selected to achieve acceptable
tuning and to stay within the emission cur-
rent range recommended by the manufac-
turer. For maximum dynamic range, the
electron multiplier voltage was set to per-
mit analytes to be detected without satura-
tion of  the multiplier at concentrations up
to four times the internal standard (IS) con-
centration of 50 ng/L specified by Method
8240.
  The  system  met all daily performance
criteria specified by Method 8240. In addi-
tion to BFB tuning criteria, these criteria in-
clude (1) minimum  RF of 0.30 for each of
the five system performance check com-
pounds (chlorobenzene, chloromethane,
1,1-dichloroethane,  1,1,2,2-tetrachloro-
ethane, and tribromomethane); (2) RF dif-
ference of <25°/o  for the six calibration
check compounds (chloroform, 1,1-dichlor-
oethene,  1,2-dichloropropane,  ethyl-
benzene, toluene, and vinyl chloride); (3) IS
retention time changes of < 30 sec; and (4)
IS peak area changes  of <50°/o.

Method Range Studies
  In method range studies, 5-mL aliquots
of reagent water were  spiked with com-
posite spiking solutions to achieve 13 con-
centrations ranging from 0.1 to 550 Ig/L for
most analytes, but 15 analytes expected to
be poorly purged were spiked at  10-fold
higher  concentrations. Eight  replicate
samples were analyzed for each of the 13
spike levels.
  A reverse library search of the data was
performed using a project-specific mass
spectral library containing the retention time
and quantitation ion of each analyte and IS.
The quantitation ion was chosen for max-
imum sensitivity while attempting to avoid
interferences from coeluting materials. For
the majority of analytes, the quantitation ion
selected was the base  peak. For Method
8240 analytes, the primary ion specified in
Method 8240 was used as the quantitation
ion. When the quantitation ion was detected
above the background, an RF was calcu-
lated  for  each analyte using the area
responses and concentrations of the
analyte and  the appropriate IS.
  To determine an estimated detection limit
(EDL) for each analyte, a trained analyst in-
spected the mass spectrum from one of the
replicate samples at the lowest concentra-
tion at which the computer detected the
quantitation  ion in at least four replicates.
The analyst examined extracted ion current
profiles of 2-5  major ions, including the
quantitation ion,  selected  from  the
reference mass spectrum. The analyte was
considered to be present if the major ions
comaximized and had  relative  intensities
within 20% of those in the reference mass
spectrum  (as specified in Method 8240),
and if  the quantitation ion gave an area
response greater than  1000 or a signal-to-
noise ratio of at least 3:1. If, in the opinion
of the analyst, the mass spectrum indicated
the presence of the analyte in question, that
concentration  was   considered  the
estimated detection limit. If the mass spec-
trum did not indicate the presence of the
analyte,  the  inspection  process  was
repeated at the next higher concentration.
   Mean RFs and  RSDs of measured RFs
were calculated at each concentration as
the first step in determining the linear range
of the method for a given analyte. The high
concentration   data were  evaluated  for
system saturation by plotting and visually
evaluating the RF as a function of analyte
concentration.  When   an  RF  obviously
decreased with increasing concentration,
appropriate concentrations were eliminatec
from the linear range. For each analyte ar
overall average RF was calculated usinc
RFs from all concentrations other than re
jected high concentrations. If the RSD foi
an overall average RF was > 25%  (an ac
ceptable  threshold  value selected with
USEPA personnel concurrence),  the con
centration range was  narrowed  in ar
attempt to achieve <25% RSDA. A concen
tration range was, however, never reduce<
to less than one order of magnitude.
  The lowest concentration at which the
analyte was identified and measured in a
least four of eight replicates was considerec
the EDL. Data obtained at the EDL were
used  to  calculate the  method  detectior
limits (MDLs) using the  USEPA procedure
described in Appendix  B to 40 CFR Par
136 (Federal Register 49 198, October 26
1984).

Matrix  Validation Studies
  Two simulated liquid waste samples wer<
prepared  for further method evaluation
One  sample  was  a  municipal  sludge
leachate prepared using a modification o
the USEPA toxicity characteristic leachim
procedure (Federal  Register 57,  21685
June 13,  1986). The other sample  was ar
artificial ground water prepared by spikm<
reagent water  with fulvic acid (Suwannei
Stream Reference, U.S.  Geological  Survey
International Humic Substance Society) a
a concentration of 1  mg/L.
  Eight replicates at each of two  analyti
spike concentrations (20 and 200 ng/L fo
most analytes; 200 and 2000 /ig/L for thi
poorly purged analytes)  and eight unspike<
replicates were analyzed for each of the twi
samples. Calibration standards prepared b;
spiking reagent water with each  analyte a
a concentration of 50 ^.g/L for most analyte!
and 500 /ig/L for the poorly purged analyte:
were analyzed at the beginning,  middle
and end of  each day.  Measured  analyti
concentrations were  calculated  using dai
ly average RFs. Precision (RSD) and ac
curacy (bias) of measured concentration:
were calculated for each analyte  at eacl
spiking concentration in each sample.

Results and Discussion

Method Range Studies

   For three compounds (acetone, 2-chlorc
1,3-butadiene,  and trichlorofluoromethane]
RSDs of measured RFs were > 25%, evei
with a concentration  range of one  order c
magnitude.  Acetone  might  yield  mon
reliable data using m/z 58 as the quantita
tion ion  rather than  m/z 43,  which  i
specified by Method 8240. A decreasing Rl

-------
with increasing concentration was evident
for 2-chloro-1,3-butadiene, which is known
to polymerize readily. The degree of poly-
merization, which would result in loss the
monomer, would be expected to be higher
at higher concentrations and could account
for lower observed RFs at higher concen-
trations. Trichlorofluoromethane was par-
ticularly sensitive to the effects of methanol
and water on GC peak shape. Other early
eluting  compounds such  as  dichlorodi-
fluoromethane (20% RSD), chloromethane
(23% RSD), and vinyl chloride (20% RSD)
also produced average RFs that were less
precise than those of most other analytes.
The range for a fourth compound  (hex-
achloropropene) that produced an average
RF with >25% RSD was  not narrowed
because the greatest  deviation from the
average RF occurred at 300 /tg/L, near the
middle  of the concentration range.

  Of the 74 analytes having a satisfactory
method range and average  RF precision,
the linear range for 56 was at least two
orders of magnitude and for 12 others was
at least 1.5 orders of magnitude. For the re-
maining six analytes, the linear range was
only one order of magnitude. Three of those
analytes (dichlorodifluoromethane, chloro-
methane, and chloroethane) were highly
volatile and two (trans-1,4-dichloro-2-butene
and  hexachloropropene)  were  poorly
purged.

  The linear range, EDL, and MDL for each
analyte  are given in Table 1. For all but 10
analytes, the EDL (the lowest concentration
at which the analyte  was  detected and
quantified in at least four of eight replicates)
was the same as the lowest  concentration
in the linear range. Lower EDLs could un-
doubtedly have been  achieved for most
analytes if MS operating conditions had
been selected to achieve maximum sen-
sitivity instead of a wide dynamic range and
measurement of high  anatyte concentra-
tions.

  For  all  analytes  except  acetone,
calculated MDLs were  considerably lower
than EDLs. For all but seven analytes each
calculated MDL was even lower, usually by
a factor of two to five, than a concentration
at which the analyte could be detected ex-
perimentally. In all cases, the highest con-
centration at which an analyte was not
detected in any of eight replicates (Table 1)
was  within  a  factor  of  three  of the
associated  EDL.  Low  calculated MDLs
reflect excellent measurement precision at
the  EDL  rather than  excessively  high
signal-to-noise ratios. The data indicate that
calculated MDLs may  be misleading.
 Matrix Validation  Studies
   The 74 analytes studied included 29 of
 the 30 compounds listed in Method 8240
 Table 6, which specifies acceptance criteria
 for data obtained from analysis of a quality
 control check sample. For 28 of those 29
 compounds,  Method  8240 acceptance
 criteria were achieved in both matrices at
 both high and low concentrations. The one
 exception was ethylbenzene spiked at the
 high concentration into reagent water con-
 taining fulvic acid; a bias of +77% was
 observed  while +62% is acceptable.
   The acceptability of measured concen-
 trations for all 74 analytes was evaluated by
 selecting a bias of +30% as an acceptance
 limit. (This limit is much more stringent that
 Method 8240  analyte-specific criteria,
 which are generally +50% or greater.) With
 a +30% bias limit, measured concentra-
 tions  were acceptable for 61  of  the 74
 analytes  spiked into  the POTW  sludge
 leachate at the high concentration and for
 63 analytes at the low concentration. Ac-
 ceptable concentrations were measured for
 50 analytes added to the fulvic acid spiked
 water at the high concentration and for 70
 at the low concentration.

   In nearly 90% of the cases in which the
 bias  of measured  concentrations  was
 >30%, the bias was positive rather than
 negative. A possible explanation of the high
 positive biases is an increased MS sen-
 sitivity when   ion  concentrations  are
 unusually  high. This effect would  be ex-
 pected to be much more noticeable when
 a capillary column is used rather  than a
• packed column, because a capillary col-
 umn produces much sharper peaks and
 higher momentary analyte concentrations
 than a packed column. The high positive
 bias was more prevalent at the  high spike
 concentration than at the low spike concen-
 tration, especially for the fulvic acid spiked
 water. The high spike concentration of 200
 i^g/L provides 1000 ng of analyte in the 5
 ml_ of sample purged. The increased sen-
 sitivity at high  concentrations was not as
 apparent in the method range study as in
 the  matrix  validation  study, possibly
 because the ion source had been cleaned
 immediately before beginning the method
 range study. Decreased source cleanliness
 may enhance the effect.

   For all but two of the cases in which the
 bias was > -30%, the low spike concen-
 tration was involved and the analyte was
 one of the 14 poorly purged analytes spi-
 ked at a 10-fold higher concentration than
 other  analytes. For those analytes, the
 calibration standard provided  2500 ng,
which could have produced a high ion con-
centration and high RF; that could account
for a negative bias at the low spike concen-
tration.  Thus,  an increased  sensitivity
resulting from high ion concentrations can
account for essentially all biases, high and
low. No evidence for a matrix effect, which
would be expected to give a negative bias,
was  observed.


Conclusions and
Recommendations

  The following conclusions and recom-
mendations are based on the results of this
study:

  • The use of  methanol as a solvent in-
   terferes with the chromatographic per-
   formance of a nonpolar capillary col-
   umn for  the determination of polar
   volatile compounds such as acetoni-
   trile, isobutyl alcohol,  and propargyl
   alcohol.
  • Methanol and water desorbed from a
   trap containing Tenax, silica gel, and
   charcoal, interfere with the chromato-
   graphic performance of a nonpolar
   capillary column for the determination
   of gaseous  and very low boiling non-
   polar compounds by a PTD procedure.
  • Of the 84 volatile compounds studied,
   74 can be determined satisfactorily by
   SW-846 Method 8240 using a VOCOL
   capillary column.
  • With MS conditions that permit Method
   8240 performance criteria to be met
   using  a capillary column and 250 ng
   of IS, an increased sensitivity may be
   observed at high ion concentrations in
   the mass spectrometer source. The ef-
   fect of MS source tuning parameters
   and cleanliness on changes in RF with
   concentration should  be evaluated.
  • Calculated MDLs en be considerably
   lower than the lowest concentrations at
   which  analytes  can  be detected
   experimentally.
  • Cryofocusing or other means to focus
   early eluting compounds to minimize
   peak broading and improve quantia-
   tion, especially at low concentrations,
   should be investigated.
  • A non-volatile,  water soluble solvent
   should be used for spiking solutions to
   avoid deleterious chromatographic ef-
   fects of methanol on early eluting
   analytes.
  • Differences between  calculated and
   observed detection limits should be in-
   vestigated to establish a protocol for
   obtaining more meaningful MDLs.

-------
TABLE 1. Linear Range and Detection Limits Obtained From Method Range Study
Applicable Experimentally
Linear Range, Determined
Analyte ^g/L EDL, ^g/La
1. Acetone
2. Acrolein
3. Acrylonitrile
4. Ally! chloride
5. Benzene
6. Bis-(2-chloroethyl) ether
7. Bromobenzene
8. Bromodichloromethane
9. Bromomethane
10. 2-Butanone
11. n-Butylbenzene
12. sec-Butylbenzene
13. tert-Butylbenzene
14. Carbon disulfide
15. Carbon tetrachlonde
16. Chlorobenzene
17. 2-Chloro-1 ,3-butadiene
18. Chlorodibromomethane
19. Chloroethane
20. 2-Chloroethyl ethyl ether
21. Chloroform
22. 1-Chlorohexane
23. Chloromethane
24. 2-Chlorotoluene
25. 4-Chlorotoluene
26. 1 ,2-Dibromo-3-chloropropane
27. 1 ,2-Dibromoethane
28. Dibromomethane
29. 1 ,2-Dichlorobenzene
30. 1,3-Dichlorobenzene
31 . 1 ,4-Dichlorobenzene
32. trans-1 ,4-Dichloro-2-butene
33. Dichlorodifluoromethane
34. 1,1-Dichloroethane
35. 1,2-Dichloroethane
36. 1,1-Dichloroethene
37. cis-1 ,2-Dichloroethene
38. trans-1 ,2-Dichloroethene
39. Dichloromethane
40. 1 ,2-Dichloropropane
41 . 1 ,3-Dichloropropane
42. 1,1-Dichloropropene
43. cis-1 ,3-Dichloropropene
44. trans-1 ,3-Dichloropropene
45. 1 ,2-Dimethylbenzene
46. 1 ,4-Dimethylbenzene
47. Ethyl methacrylate
48. Ethylbenzene
49. Hexachlorobutadiene
50. Hexachloroethane
51. Hexachloropropene
52. 2-Hexanone
53. lodomethane
54. Isopropylbenzene
55. p-lsopropyltoluene
56. Methacrylonitrile
57. Methyl methacrylate
58. 4-Methyl-2-pentanone
59. Naphthalene
60. Propionitrile
61. n-Propylbenzene
62. Styrene
63. 1,1,1,2-Tetrachloroethane
64. 1,1,2,2-Tetrachloroethane
65. Tetrachloroethene
5500-170
5500-100
5500-170
550-10
550-3.0
5500-55
170-3.0
550-3.0
550-5.5
5500-100
170-1.0
170-1.0
170-1.0
550-3 0
550-3.0
300-3.0
550-5.5
550-3.0
550-30
5500-30
550-3.0
550-3.0
550-30
550-3 0
550-3.0
550-10
550-3.0
550-3 0
300-3 0
300-3 0
300-3.0
550-170
55-5.5
550-5 5
550-5.5
550-5.5
550-5.5
550-5.5
550-5 5
550-5.5
550-3 0
550-5.5
550-5.5
550-5.5
550-1.0
550-10
1000-10
550-1.0
300-3.0
550-5.5
3000-170
5500-30
550-5.5
300-3.0
300-10
5500-30
3000-30
1700-30
300-3.0
5500-170
170-1.0
300-3.0
300-3.0
550-5.5
300-3.0
170
55
170
5.5
3.0
55
3.0
3.0
5.5
30
1.0
1.0
1.0
30
30
30
5.5
3.0
17
30
3.0
3.0
17
30
30
10
3.0
30
30
30
30
55
55
5.5
55
5.5
5.5
55
5.5
5.5
30
55
55
5.5
1.0
1.0
10
1.0
3.0
5.5
170
30
5.5
3.0
10
30
30
30
3.0
170
1.0
3.0
3.0
5.5
3.0
Calculated Nondetection
MDL, tg/L" Limit, ^g/Lc
200
10
40
2
0.2
10
0.7
02
2
10
0.4
0.4
0.8
0.3
0.2
0.3
2
0.2
5
9
01
0.1
10
0.3
03
2
0.2
0.2
0.5
0.4
0.5
30
1
0.7
0.4
1
1
0.7
0.8
0.7
0.3
0.5
0.9
1
0.2
0.6
7
04
0.7
2
50
6
2
0.4
0.7
9
5
6
0.5
40
0,2
0.4
0.2
1
0.2
100
30
100
3.0
1 0
30
1.0
1.0
3.0
10
0.3
0.3
0.3
1.0
1.0
1.0
3.0
1.0
10
10
1.0
1 0
10
1 0
1.0
5.5
1.0
1 0
1 0
1 0
1.0
30
30
3.0
3.0
3.0
3.0
30
30
3.0
1 0
30
30
3.0
03
0.3
30
0.3
1.0
3.0
100
17
3.0
0.3
5.5
17
3.0
17
1.0
100
0.3
1 0
1.0
3.0
1.0

-------
TABLE 1.
Continued
Analyte
66. Toluene
67. Tribromomethane
68. 1 ,2,4-Trichlorobenzene
69. 1,1,1-Trichloroethane
70. 1.1,2-Trichloroethane
71. Trichloroethene
72. Trichlorofluoromethane
73. 1,2,3-Trichloropropane
74. 1,2,4-Trimethylbenzene
75. 1 ,3,5-Trimethylbenzene
76. Vinyl acetate
77. Vinyl chloride
Applicable
Linear Range,
iig/L
550-3.0
300-3.0
300-3.0
550-3.0
550-5.5
550-3.0
300-30
550-17
300-3.0
300-3.0
3000-55
550-5.5
Experimentally
Determined
EDL, ng/L"
3.0
3.0
3.0
3.0
5.5
3.0
30
17
3.0
3.0
55
5.5
Calculated Nondetection
MDL, ng/L" Limit, pg/Lc
0.2
0.3
0.3
0.2
0.3
0.1
10
2
1
0.3
4
0.6
0.3
1.0
1.0
1.0
3.0
0.3
17
10
1.0
1.0
30
3.0
"Experimentally determined estimated detection limit.
"Calculated method detection limit.
cNondetection limit is the highest concentration studied at which the analyte was not detected.
  M. F. Yancey, R. A. Kornfeld, andJ. S. Warner are withBattelle—Columbus Division.
    Columbus, OH 43201-2693.
  Thomas Pressley is the EPA Project Officer (see below).
  The  complete  report,  entitled  "Capillary Column GC-MS Determination of 77
    Purgeable Organic Compounds in Two Simulated Liquid Wastes," (Order No.
    PB 88-245 881/AS; Cost: $14.95,  subject to change) will be available only
    from:
          National Technical Information Service
          5285 Port Royal Road
          Springfield. VA 22161
          Telephone: 703-487-4650
  The EPA Project Officer can be contacted at:
          Environmental Monitoring and Support Laboratory
          U.S. Environmental Protection Agency
          Cincinnati, OH 45268

-------
                                                                                                               U
                                                                                                    oa;ncs i-
United States
Environmental Protection
Agency
Center for Environmental Research
Information
Cincinnati OH 45268
                                                                                                             BULK RATE
                                                                                                        POSTAGE & FEES
                                                                                                          PERMIT No. G-3
Official Business
Penalty for Private Use $300

 EPA/600/S4-88/030
                          OOOC329    PS

-------