United States
Environmental Protection
Agency
Environmental Monitoring
Systems Laboratory
Research Triangle Park NC 27711
Research and Development
EPA-600/S4-84-025  May 1984
Project  Summary
A  New  Audit Method for  EPA
Reference Method  6
R.K.M. Jayanty
  A simple, inexpensive, and accurate
 method for evaluating and/or auditing
 sampling and analytical phases of
 EPA Source Reference Method 6 was
 developed. The method uses a known
 amount of a chemical compound in the
 form of a tablet or pill (or placed in a
 capsule) to  generate  sulfur dioxide
 (SO2) quantitatively by reaction with an
 acid.  The reaction takes place in a
 compact glass impinger  system that
 can  be taken to  the  field.  The SOa
 generated in test runs was collected and
 analyzed using the Method 6 procedure.
 The  SO2 generation was quantitative
 and recoveries were found to be 94 ± 5
 percent. The time to complete the
 reaction was less than 15 min at a flow
 rate of 1 L/min, but the recommended
 sampling time was 45 min. The tablets
 prepared gravimetrically were found to
 be stable over a 6-month period. The
 between-laboratory results obtained
 showed close agreement with the
 expected concentrations  based on
 calculations from the stoichiometric
 reaction. The estimates of repeatability
 (or within-laboratory precision) were
 ±5.0 mg 95 percent of the time for the
 two  concentration levels tested. The
 reproducibility (or between-laboratory
 precision estimate) was within ±5.9 mg
 at the low range and ±23.8 mg at the
 high range about 95  percent of the
 time. The maximum bias observed for
 the method was 0.5 percent.
  This Project Summary was developed
 by EPA's Environmental Monitoring
 Systems Laboratory, Research Triangle
 Park. NC, to announce key findings of
 the research project that is fully docu-
 mented in a separate report of the same
 title (see Project Report  ordering
 information at back).
Introduction
  The U.S.  Environmental Protection
Agency (EPA) is interested in developing
audit  materials for all EPA source
reference methods. These audit materials
are used to conduct performance audits
during stationary source measurements.
EPA is currently providing liquid sulfate
standards as audit materials for  EPA
Source Reference  Method 6,  which is
used to determine  SOz emissions from
stationary sources. However, these liquid
sulfate audit standards are useful only for
evaluating the analytical phase of Method
6. Currently, sampling phase procedures
are evaluated only by a systems audit; i.e.,
someone observing the sampling proce-
dures in the field. Therefore, a method or
device that  can be used as  an audit
material for the evaluation of both  the
sampling and  analytical phases of  the
source reference methods is needed.
  Research Triangle Institute (RTI)  was
contracted by  the Quality Assurance
Division  of EPA's  Environmental Moni-
toring Systems Laboratory, Research
Triangle Park, North Carolina, to develop
a simple method or device for evaluating
and/or auditing both the sampling and
analytical  phases of various source
reference methods in the field. The main
objectives of this study were to develop a
device for auditing both the sampling and
analytical phases of EPA Method 6 and to
establish its accuracy, precision, and
stability. The  method  uses a known
amount of a chemical compound (sodium
bisulfite  or sodium sulfite)  mixed with
an inert binder (polyvinyl pyrollidine) and
formed into a tablet or pill to generate 862
quantitatively by reaction with sulfuric
acid. The SOz generated in test runs was
collected and analyzed using the Method
6 procedure as published in the Federal

-------
Register (Vol. 42, No. 160, August 1977).
The  audit method developed was  also
evaluated at two concentration levels by
six different laboratories to determine its
repeatability, reproducibility, and accuracy.
The  various activities performed during
the study and the results obtained are
discussed in this report.

Experimental Procedures
  The preliminary experiments to gener-
ate 862 from the chemical reaction of
sodium bisulfite and sulfuric acid were
conducted using the experimental system
shown in Figure 1. Initially, a known
amount  of sodium bisulfite was placed
in a three-way, stoppered, round-bottomed
flask and dilute sulfuric  acid (10 percent)
was  added slowly from a burette. Dry
nitrogen or air was bubbled through the
solution at a flow  rate of  1  L/min to
sweep the released SO2 into  a glass
manifold. Glass wool was used to remove
the aerosol vapors. The generated SO2
was  monitored continuously with a SO2
source analyzer (TECO Model 40 fluores-
cent analyzer) to determine the time of
completion  of  the reaction. Within 10
minutes the reaction was completed. In
later experiments, the SOz gas generated
from the chemical reaction was collected
into  an  evacuated Tedlar bag. After 30
minutes of gas flow into the bag, the
contents of the  bag were further diluted
with zero air and analyzed for SOz using
the source analyzer. The analyzer  was
calibrated with a  cylinder  gas  of  SOz
referenced  to  the National  Bureau of
Standards Standard Reference Materials.
The  experiments were repeated using
various amounts of sodium bisulfite and
different concentrations of sulfuric acid.
  The sampling train component parts,
except for the SOz gas generator, were
the same as those in the EPA Method 6
train. The sampling  probe was not
required and hence not included in the
train. A modified midget impinger  (Ace
Glass, Inc., Model #7544-35) was used
as an SOz generator in place of a test-site
stack sample to generate a gas sample
(SOz) from the chemical reaction.
  The sample collection procedure  was
followed as described in EPA Method 6
except for  the  following modifications.
The gas sample was drawn through the
SOz generator for 45 minutes. The tablet
generally took  15 minutes to dissolve
completely; however, the sample collection
was continued for an additional 30
minutes. Since ambient air was used to
draw the sample from  the generator, a
blank run (without a tablet) was performed
for 45 minutes and impinger solutions
were analyzed.
                                                               Connected to an
                                                              Evacuated Tedlar
                                                                 Bag or SO2
                                                               Monitor Intake
                            3-Way Round-
                            Bottomed Flask
Figure 1.   Preliminary SOi generation system.
  The impinger solutions were analyzed
after each run for sulfate (converted form
of SOz collected) by ion chromatography
instead of barium perchlorate-thorin
colorimetric titration as described in EPA
Method 6. However,  it was established
initially that both analytical procedures
gave identical results.
  Because the amount of sodium bisulfite
or sodium sulfite required to generate
source concentrations of SOz (—400- to
1,000-ppm levels) is  so. small, an inert
binder (polyvinyl pyrollidine,  PVP) was
used for mixing. The mixture was made in
the form of a tablet or pill. The tablets or
pills were made using  either a KBr die
(commonly used to make KBr pellets for
infrared spectrophotometric studies) or a
tablet  press. The tablets or pills  were
made by  adding to  the die a layer of
binder,  a  weighed amount  of  sodium
bisulfite, and  another  layer of binder.
Several hundred pounds of pressure was
applied using a hydraulic press  to form
the tablet. The PVP binder was found to
be inert and water-soluble and, because
the chemical compound was contained
between the two layers of binder,  the
tablets could  be handled  safely. The
binder also helped reduce the rate of
reaction of SOz generation. Tablets were
made with varying amounts of  sodium
bisulfite to generate different concentra-
tions of SO2.


Collaborative Testing
Procedures
  From those laboratories having contracts
with EPA, six were selected for the
collaborative study based on their past
experience in source  testing. They were
Engineering-Science,  Inc., Entropy Envi-
ronmentalists, Inc., PEDCo Environmen-
tal, Inc., Radian Corporation, TRC-Envi-
ronmental Consultants, Inc., and TRW-
Environmental Engineering Division.
Seven tablets representing two concen-
tration levels were shipped to each of the
six laboratories for analyses. Of the seven
tablets, three were low concentration (60
mg SO2) and four were high concentra-
tion (120 mg SO2). The true concentra-
tions of the tablets were based on
theoretical concentrations calculated
from gravimetric preparations and  stoi-
chiometric chemical  reaction. A  SOz
generator (modified midget impinger),
general instructions to perform the
experiments, and  a data sheet for
reporting the results were  also sent to
each  laboratory.  Each laboratory  was
provided with the true concentration for
one tablet and was asked to analyze the
true concentration tablet  first; if the
measured value were within ±10 percent,
they were to  analyze the remaining six
tablets. The  trial  run  was  thought
necessary because most laboratories had
not used this method before. The labora-
tories were informed that these samples
were  not for an audit but  were for an
evaluation of the developed method. The
types of compounds in the tablets were
not disclosed.

Results and Discussion
  The measured SOz concentrations
from the preliminary Tedlar bag experi-
ments and the corresponding  expected
concentrations from the  stoichiometric
chemical reaction are shown in Table 1.
The results  show  that  the percent
difference obtained ranged from 20 to 25
percent low between the concentration of
SOz measured and  the  expected  SOz

-------
concentration from the stoichiometric
reaction for various amounts of sodium
bisulfite. The SOz generation was also
found to be independent of flow rate
(varied 0. 1 to 1 .4 L/min) and the amount
of sulfuric acid. The sulfuric acid must be
slightly in excess of the amount of
sulfuric acid. The sulfuric acid must be
slightly in excess of the amount actually
required for completion of the reaction.
The lower measured SO: concentrations
were probably a result of the combination
of loss of SOz in the bag plus loss of wet
S02 on the glass system, but the actual
reasons are not known nor were they
determined. However, the percent differ-
ence was constant within experimental
error for various amounts of sodium
bisulfite.
The results obtained for different
amounts of sodium bisulfite (ranging
from 10 to 800 mg) using the described
method are shown in Table 2. As the
results indicate, the percent difference
between the measured concentration
and the expected concentration ranged
from 2 to 7 percent; i.e., SO2 recoveries
obtained were between 93 and 98
percent. Lower SOz recoveries were
obtained for small amounts of sodium
bisulfite (10- to 40-mg range). Similar
results were obtained for sodium sulfite.
However, the S02 recoveries were found
to be poor for sodium metabisulfite. Since
ambient air was used to draw the sample
from the generator, a blank run (without a
tablet) was performed to test the ambient
air. Analysis results showed a negligible
amount of SOz. The SOz generated from
the material was found to be independent
of flow rate and total volume of sample
collection.
A study was also performed to determ ine
the storage stability of the prepared
tablets over a 6-month period at room
temperature. The prepared tablets were
analyzed periodically. The results of this
study are shown in Table 3. These results
demonstrate that the tablets are stable
within experimental error over the 6-
month period. The stability study is still in
progress.
The results of the analyses of the
samples by the different laboratories and
the expected concentrations are shown
in Table 4. The analysis results, in
general, showed close agreement between
the expected values. For example, results
of analyses of the low concentration
tablets ranged from -4 to +7 percent
different from the expected value. Like-
wise, the results of analyses of the high
concentration tablets ranged from -5 to
+8 percent different from the expected
Table 1. Bag Results
Amount of Amount of SOz Amount of SOz
rVaHSOs expected" measured
(mg) Ippm) (ppm)
102.6 650 507
199.9 804 646
2OO.5 863 66O
2OO.6 8O1 604
202.9 869 676
2O1.4 799 652
207.5 1O03 794
2O2.3 842 • 712
202.5 739 593
4O1.9 944 • 721
"Calculated from stoichiometric chemical reaction.
*Prm>nt rtiffevt-nff = lOO>r Measured SOz concentrate - Expected SOz concentrate
Expected SOz concentrate
Table 2. Results of SOz Generated from the Prepared Tablets* b
Weight of Amount of SOz ' Amount of SOz
NaHSOa expected measured
(mg) (mgf (mg)
12.9 7.9 . 5.2
4O.1 24.6 21.9
50.6 31.1 31.3
10O.4 61.8 57.6
101.6 62.5 61.5
102.9 63.3 59.6
202.9 124.9 120.3
401.9 247.3 235. 1
802.3 493.7 464.4
114.4" 38.5 23.9
204.3' 103.8 105. 1
'Flow rate 1 L/min for 45 minutes.
"3 to 5 mL of 10 percent HzSO* added.
^Calculated from stoichiometric chemical reaction.
"Percent riiff prince =inn * Amount of SOz measured - Amount of SOz expected
Amount of SOz expected
"Sodium metabisulfite was used.
'Sodium sulfite was used.
Table3. Stability Study Results
Amount of SOz Amount of SOi
No. of expected measured
days (mg) (mg)'
0 122.7 121.9
15 123. 1 125.5
29 123.0 124.9
48 123.2 126.8
68 121.7 124*
94 122. 1 126S
180 121.5 125.3
'Percent difference - 10O ..Amount ^oOzmeasured- Amount of SOzexpected
Percent
difference"
-21.9
-19.7
-23.5
-24.5
-22.2
-18.4
-20.8
-15.4
-19.8
-23.6
Percent
difference"
-34.2
-11.0
+0.6
-6.8
-2.0
-5.8
-3.7
-2.0
-5.9
-37.9
+1.3
Percent
difference"
-0.65
+1.95
+ 1.54
+2.90
+2.20
+3.55
+3.13

Amount of SOz expected
value. Two values reported from two Conclusions
laboratories were discarded during the
statistical analysis because the labora- A new audit method that is simple and
lories identified problems during the SOz inexpensive was developed for evaluating
generation. Blank values reported by the both the sampling and analytical phases
laboratories were also negligible. of EPA Method 6 in the field. The SOz

-------
   generation was found to be quantitative
   and the recoveries were found to be 94 ±
   5 percent over the 40- to 800-mg range of
   sodium bisulfite. The  method was also
   evaluated at two concentration levels
   with six different laboratories to determine
   its repeatability, reproducibility, and bias.
   The  interlaboratory  results obtained
   showed close agreement with the expected
   concentrations based on calculation from
   the stoichiometric reaction. The repeata-
   bility within a laboratory was found to be
   within ±5.0 mg 95 percent of the time for
   the two concentration levels tested. The
   reproducibility between laboratories was
   within  ±5.9 mg at the  low range and
   ±23.8 mg at the high range about 95
   percent of the time. The maximum bias
   observed  was 0.5 percent by comparison
   with the  expected concentrations. The
   SOa  generation from  the material was
   found to be independent of flow rate and
   total volume, and  the prepared  tablets
   were stable at room temperature over  a
   6-month period. It is recommended that a
   performance audit using this new audit
   method for EPA Method 6 be used during
   future source SC>2 emissions analyses as
   a routine quality assurance procedure.
       Table 4.    Interlaboratory Results (mg)

Lab 1
RV
EV
Lab 2
RV
EV
Lab 3
RV
EV
Lab 4
RV
EV
Lab 5
RV
EV
Lab 6
RV
EV
Known

119.7
121.8

126.5
122.2

114.2
122.2

123.2
122.7

116.1
121.8

116
122.4

1

59.9
60.3

64.5
60.3

a
-

63.0
60.5

60.1
59.9

58.3
60.9
Low
2

59.1
60.4

61.5
60.2

58.2
60.7

62.3
604

60.7
60.0

a
-

3

60.4
60.3

57.9
60.4

58.1
60.7

61.9
60.5

62.0
603

60.2
60.8

1

119.0
122.2

129.4
122.0

118.7
121.9

128.4
122.4

1167
122.1

117.0
122.4
High
2

119.9
122.3

128.0
121.8

115.8
122.1

124.9
122.6

119.6
122.1

118.0
122.3

3

122.6
122.1

131.7
121.9

115.6
122.0

1256
1225

120.5
122.2

118.0
122.3
       RV - Report value.
       EV = Expected value.
       "Values discarded.
          R. K. M. Jayanty is with Research Triangle Institute, Research Triangle Park, NC
            27709.
          Robert G. Fuerst is the EPA Task Manager (see below).
          The complete report, entitled"A New Audit Method for EPA Reference Method 6,"
            (Order No. PB84-172 097; Cost: $8.50, subject to change) will be available only
            from:
                 National Technical Information Service
                 5285 Port Royal Road
                 Springfield, VA 22161
                  Telephone: 703-487-4650
          The EPA Task Manager can be contacted at:
                 Environmental Monitoring Systems Laboratory
                 U.S. Environmental Protection Agency
                 Research Triangle Park, NC 27711
                                                                                 U S GOVERNMENT PRINTING OFFICE. 1984 — 759-015/7688
urn*-.i states
Environmental '
Agency
Center for Environmental Research
Information
Cincinnati OH 45268
Official Business
Penalty for Private Use $300
       CHlCAUU  IL

-------