EPA-600/S4-84-031  May 1984
Project Summary
Study  of  Carbon  Monoxide
Exposure  of  Residents  of
Washington,  DC  and
Denver,  Colorado
T. D. Hartwell, C. A. Clayton, R. M. Michie, R. W. Whitmore, H. S. Zelon, S. M.
Jones, and D. A. Whitehurst
  This report describes a study funded
by the EPA and conducted by the
Research Triangle Institute in 1982 and
1983 to evaluate methodology for
collecting  representative  personal ex-
posure monitoring (PEM) CO and cor-
responding activity data in an urbanized
area. This involved telephone screening
of households and sample selection of
respondents in the metropolitan areas
in and around Denver, Colorado and
Washington, D.C. Data on CO  breath
levels were also collected in Washing-
ton, D.C. (PEDCo Environmental con-
ducted the field work in Denver.) The
target population in both cities consist-
ed of the non-institutionalized, non-
smoking adults (ages 18 to 70) of these
metropolitan areas. The data collected
in the field were edited and appropriate-
ly weighted to produce CO exposure
estimates for the target population.
  Estimates of CO  exposure for the
winter of 1982-83 in Washington, D.C.
were obtained using  the data base
constructed from the raw CO levels by
activity data. This data was collected
over a 24-hour period when the respond-
ent carried a CO PEM and an activity
diary. The data consisted of hourly CO
values on 712 respondents, activity
patterns and corresponding CO levels
on 706 respondents, and CO  breath
measurements corresponding to the
PEM CO data on 669 respondents. The
size of the target population was esti-
mated to be 1.22 million individuals.
  The weighted average maximum hour-
ly PEM CO level in Washington, D.C.
was 6.74 ppm. The average maximum
8-hour CO level was 2.79 ppm. The
percentage of the population with maxi-
mum hourly CO values over the 35 ppm
CO standard was estimated to be 1.28
percent while the percentage with an
8-hour maximum over the 9 ppm stand-
ard was 3.9 percent.
  Estimates were also made for sub-
groups of the population. Persons in
high-exposure occupations (about 4.6%
of the total population) generally exhib-
ited higher CO exposure levels: it was
estimated that about 24% of this high-
exposure group had 1 -hour CO expo-
sures above the 36 ppm standard and
that about  28% exceeded the 8-hour
standard. It was also shown that CO
levels were generally higher for com-
muters, especially for those with larger
amounts of travel.
  By combining PEM data with data
from individuals' diaries, estimates of
both CO levels and time durations for
various activities and personal environ-
ments were made. For example, the
activities "in parking garage or parking
lot" and "travel, transit" had the high-
est average CO concentrations (6.93
ppm, and 4.61 ppm, respectively) while
"sleeping"  had an estimated CO con-
centration of only .86 ppm.
  Variation from duplicate  hourly PEM
measurements under field condition
were also analyzed. An  analysis of
variance of this data which considered
person-to-person,  hour-to-hour, and
measurement variation indicated that
about 6 to 6 percent of total variation

-------
among the hourly duplicate readings
was due to deviation in measurements
made by two PEMsatthe same hour for
the same person.
  This Project Summary was developed
by  EPA's  Environmental Monitoring
Systems Laboratory, Research Triangle
Park. NC, to announce key findings of
the research project that is fully docu-
mented in a separate report of the same
title (see Project Report ordering infor-
mation at back).
Introduction
  As the control of emission increases,
the burden of proof on EPA to show that a
particular level of emission control is
justified also increases. It has become
more and more important to show that a
given level of control is justified for each
air pollutant, with the relative risk of
public health approximately comparable
for each pollutant controlled.
  A critical  factor in determining the
degree of risk to the population is the
exposure of members of the population.
In the past, monitoring of airborne pollut-
ants has necessarily been based on the
assumption that fixed-site monitoring is
representative of concentrations sur-
rounding the site, since  monitoring tech-
niques were generally not developed for
determining personal exposures. Then to
obtain estimates of population exposure,
techniques such as computer simulation
or overlaying isopleths of pollution con-
centrations measured at fixed  sites on
population density maps have been used.
For some pollutants, these techniques
may be reasonable approximations; how-
ever, recent work has shown that many
pollutant concentrations are  not homo-
geneous and that activity patterns play an
important role in an  individual's actual
exposure. Therefore, data from ambient
fixed sites often differ significantly from
the concentrations with  which  people
actually come into contact.
  Accordingly, RTI and EPA formulated a
study plan to  develop  and field test a
population exposure methodology using
CO while making sure that the method-
ology was broad enough to accom modate
other pollutants of concern. The specific
objectives of this study  were the follow-
ing:
  —  To develop a methodology for meas-
     uring the  distribution of carbon
     monoxide (CO) exposures of a repre-
     sentative population of an urban
     area for  assessment of the risk to
     the population.
 —  To test, evaluate, and validate this
     methodology by employing it in the
     execution of pilot field studies in
     Denver, Colorado, and in Washing-
     ton, D.C.
 —  To obtain an activity-pattern data
     base related to CO exposures.
  Carbon  monoxide was selected for
primary emphasis in the current study
because:
 —  Accurate and portable field-tested
     instruments now are available for
     CO.
 —  Most of the CO to which the public
     is exposed  can be attributed to
     motor vehicles.
 —  It appears that CO is a good "indi-
     cator" (i.e., surrogate) pollutant for
     estimating  exposures  to  several
     other motor vehicle pollutants of
     interest.
 —  Because CO is a nonreactive air
     pollutant,  it is simpler  to treat
     analytically.
 —  The health effects of CO are reason-
     ably well documented, and NAAQS
     based  on these effects have been
     promulgated.
 —  Considerable data exist  showing
     that CO varies spatially and  that
     many locations in cities have con-
     centrations  that differ from those
     reported at fixed air monitoring
     stations.
  The study was carried out in Washing-
ton, D.C. and Denver, Colorado during the
winter of 1982-83 (the period of the year
with maximum ambient CO concentra-
tions).  The population exposure profile
was determined by direct measurement
of  CO with personal exposure monitors
(PEMs) through the use of statistical
inference  from the statistically drawn
sample. The study  provided sufficient
data to determine exposure as a function
of  concentrations  within  significant
microenvironments (home,  in-transit,
work, and leisure) and individual activity
patterns.
  The  report describes  in detail  the
activities, results,  and recommendations
evolving from the study. It is extremely
important to note that the study not only
developed and tested methodology for
measuring the distribution of CO in an
urban  area  but also produced direct
estimates of CO exposure that apply to
two large metropolitan areas. In addition,
a very important product of this work is a
unique and valuable data base on indi-
vidual exposures  to CO and the corre-
sponding  activities that led to these
exposures.
Summary of Study  Design and  4
Procedures                         "

  The target population consisted of the
non-institutionalized, non-smoking adults
(ages 18 to 70) in the metropolitan areas
in and around Denver,  Colorado and
Washington, D.C. A probability sample of
the target population was selected in both
cities. This sample was a stratified, three-
stage,  probability-based  design.  Area
sample  segments defined by Census
geographic variables were selected at the
first stage of sampling. Households were
selected at the second stage, and a house-
hold member was administered a short
screening interview covering all house-
hold members to identify individuals with
characteristics believed to be positively
correlated with CO exposure. Thus, house-
hold members with these characteristics
could be oversampled in the third stage.
Donnelley Market Corporation listings
were used to help select households for
the screening interview. The third stage
sample was  a stratified sample  of
screened eligible individuals (i.e., non-
smoking, aged 18 to 70). The individuals
in the third stage sample were admin-
istered a Computer Model Input Ques-
tionnaire  and were  asked  to carry a
personal CO monitor and an Activity  |
Diary for 24 or 48 hours (for Washington  *
and Denver, respectively). A breath sam-
ple was also requested from these indi-
viduals and they were asked to fill out a
Household (Study) Questionnaire. The
third stage sample design also allocated
individuals to  specific days  within the
sampling period. A detailed discussion of
the sample design is given in the report.
  To carry out the sample design, RTI
developed the data  collection instru-
ments and worked with EPA in obtaining
OMB approval  for the study. An  initial
telephone screening was carried out in
both Denver  and Washington,  D.C.  by
using RTI's Computer Assisted Telephone
Interviewing (CATI) system.  This  initial
screening was supplemented by limited
field screening in both  sites. Specific
information  collected  during this  inter-
view included: time  spent in regular
commuting and smoking  status of each
household member, as well as presence
of gas appliances and attached garages in
their residences. After the initial screen-
ing and the initial  selection of potential
participants, another telephone interview
was conducted. The purpose of this call
was to contact the selected individual to
further explain the study  and attempt to
enroll  him (her) into  the study.  If the  M
individual agreed to be part of the study,  ™

-------
an appointment was established for a
field interview. In addition, during this
call, a Computer Model Input Question-
naire was administered which collected
additional data on commuting patterns,
demographics of household  members,
and household characteristics.
   Finally, participating individuals were
met at their home or other convenient
location  and  given all study  materials.
These participants carried both a PEM (a
model  COED-1, which  utilized a data
acquisition package supplied by Magnus,
Inc.) for the 24 hours of their participation
and an Activity Diary  in which to record a
description of their activities. Participants
were requested to push a button on their
PEM every time they changed activities
and to record descriptions of the new
activities in their diaries. In addition, for a
small sample of participants, a GE/HP
PEM (which  utilized  a Hewlett-Packard
HP-41CV programmable calculator) was
used  which allowed  the participant  to
also enter an activity code into the
monitor. Participants were also asked to
complete a self-administered Household
Questionnaire which provided informa-
tion on themselves and on their home and
work environments. The telephone screen-
ing and  sample selection  of individuals
for both Denver and Washington were
carried out by RTI as was the field work in
Washington.
  The results of the telephone screening
and field activities for  the  study are
described in detail in  the report. Briefly,
8643  household  screenings  were at-
tempted by RTI in Washington, D.C. and
4987 were attempted in Denver,  Colo-
rado.  The  successful screening  rates
were 75.8  percent in Washington and
70.4 percent  in  Denver.  From  these
telephone  and  field  screenings,  5418
eligible respondents  were identified in
Washington and 2232 in Denver. From
this population of eligibles, 1987 indi-
viduals were  selected for participation
(i.e., to carry a PEM) in Washington and
1139 in Denver. Of these selected indi-
viduals, 58 percent actually scheduled
appointments to carry a PEM in Washing-
ton. Finally, 35.8 percent of the indi-
viduals in Washington selected to partici-
pate contributed usable CO monitor data.
This represented 712 sample  respond-
ents. Instrument failure was one of the
major reasons for the low response rate.
Specifically, CO data was not collected or
was unusable for analysis purposes for
232 respondents (22% loss rate) due to
monitor failure or malfunction. Usable
CO breath data corresponding to the
usable CO monitor data was collected on
659 sample respondents.
  In order to successfully implement the
study in Washington, D.C., a field office/
laboratory was established in the offices
of the Metropolitan DC Council of Govern-
ments. This  office was used for several
purposes  including  supervision of field
staff, storage of supplies, maintenance of
records, allocation of field assignments,
and maintenance and repair of the PEMs.
This office was visited twice nightly by all
interviewers to receive PEMs and data
collection forms for that evening and for
return of completed study materials in-
cluding the PEMs used the previous 24
hours. All calibrations of the PEMs during
the study were carried out in this field
laboratory. In addition to the field super-
visor for the interviewers, the field labora-
tory was staffed with two full-time tech-
nicians working seven days per  week
throughout the study. A detailed descrip-
tion of the PEMs (COED-1 s and GE/HPs)
used in this study and the extensive daily
technical  support that  they  required is
given in the report.
  As  mentioned above, breath samples
were  collected from respondents during
the study. This required RTI to evaluate a
method for  collecting and  measuring
alveolar CO. The method essentially
required each respondent to blow  into a
sample bag  at the end of  his 24-hour
sampling  period.  This  sealed bag was
then returned to the field laboratory for
CO analysis.
  Throughout the field work, a quality
control and assurance program  was
maintained for the sampling and analysis
procedures employed. This included using
field  standards to calibrate  all  the CO
monitors. The monitors were subject to
calibration (two-point, zero/span) before
they were put in the field and 24 hours
later when they were returned from the
field. The comparison of the two calibra-
tion curves was used to assign validity
codes to the PEM data. Other quality
control procedures employed were: a ten
percent check of data transcribed from
monitor  memory to  field  data  sheets;
monitoring control charts on each monitor
describing the course of differences be-
tween pre-sample and post-sample span,
zero,  battery  voltage, and flow rate
values; collecting duplicate colocated
samples for the purpose of characterizing
monitor precision; performing  external
and internal  QA and QC audits; perform-
ing multipoint calibrations to assess
monitor linearity during the study; and
obtaining duplicate breath samples from
respondents. The results of these exten-
sive quality control and quality assurance
procedures are given in the report.
  After the field work was completed, the
data were returned to RTI where detailed
editing of the data was carried out by RTI
editors. The  data were then entered into
computer files using RTI's mini-computer
data base entry system. All data were
keyed and then 100 percent key-verified.
Extensive machine editing was carried
out which resulted in identifying many
computer records which required further
manual editing. The process of editing the
computer files took extensive staff time.
In particular, checking the consistency of
the PEM data with the diary data was a
time consuming process.
  Sampling  weights were  computed ac-
cording  to  prescribed formulas. This
involved  extensive computations so that
the  weights could  be used to draw
inferences to the target populations. The
sampling weights  were then put on  a
computer file so  that they could be
merged with the corresponding field data.
  Detailed statistical analyses were car-
ried  out  using computer data files with
PEM CO and activity diary data. Estimates
computed during this analysis were
weighted estimates for the population of
inference—adult non-smokers  in the
Washington, D.C. metropolitan area. Stand-
ard errors of estimates were produced by
using specially written software designed
for analysis of data from complex sample
surveys.
  In  particular, analyses were first pro-
duced for hourly CO exposure data. These
analyses included  computing statistics
describing diurnal patterns,  maximum
hourly CO concentrations, maximum 8-
hour CO concentrations, and mean hourly
CO  concentrations. Statistics included
means, standard errors, and percentages
of the population exceeding certain speci-
fied CO levels. Estimates of these statis-
tics were computed for all days, week and
weekend days, and low and high CO days
(as indicated by fixed-site monitors).  In
addition, CO hourly level comparisons
were also  made  for  3 occupational
groups; 6 commuter groups  (i.e., non-
commuters;  commuters who traveled up
to 5  hours/week; etc.); and 4 categories
describing the use of gas stoves.

  Estimates were also produced for CO
exposure levels for various activities (e.g.,
in transit) and locations (e.g., indoors—at
residence). Statistics computed for each
activity and  location included mean CO
level, the estimated standard  error, and
estimates of the proportion of the popula-

-------
tion  having CO  levels above specified
levels. The distribution of times spent in
the various activities and locations were
also computed.
  Breath measurements taken at the end
of each  individuals' monitoring  period
were  used to produce estimates of the
distribution of CO breath  levels in the
Washington, D.C. area. Finally, using the
duplicate CO monitor data,  estimates
were  computed  to  assess variation  in
PEM  measurements under field condi-
tions.

Summary of Study Results and
Conclusions
  Based on the experience gained during
the Washington, D.C.  and  Denver PEM
CO studies, the methodology developed,
with some modifications (see the detailed
report), may be used effectively in other
areas of the country for collecting PEM
data.  Experience gained during this
initial study will improve the execution of
such similar studies. Modifications that
are suggested include a different  sam-
pling  design  using  the  classified tele-
phone directory listings, improvements in
the CO monitors, and additional refine-
ment of the method used to  collect
activity data. These modifications should
make the methodology more cost effec-
tive, improve the response rate, and lead
to more accurate activity information.
  Important new information was learned
for each of three sampling methodology
studies of the project: (1) It was found that
geographically classified telephone direc-
tory listings can be used in a cost-effective
manner in association with standard area
household sampling techniques for per-
sonal monitoring studies like the current
CO study. The sampling design for the
cost-effective use of these telephone
directory  listings differs  substantially
from the design  used for the  CO study,
however (details are given in the report).
(2) Sending lead letters to individuals who
were  selected for  personal monitoring
prior to calling to schedule an appoint-
ment was found to be an effective strate-
gy. (3) The need for person-day sampling
for studies that monitor personal expo-
sure to airborne pollutants is apparent.
The CO study gained valuable experience
with this technique.  Further study, possi-
bly even another methodological study, is
needed to refine this technique.
  Based  on experience derived during
this project, two important conclusions
were  reached concerning the  use of the
COED-1 and GE/HPmonitorsfor monitor-
ing personal CO exposure:
 —  The COED monitors exhibited a less
     than desirable reliability during this
     study producing a final successful
     sample completion rate of only 78
     percent.  Since most of  the  lost
     samples can be attributed to unreli-
     ability of the monitor electronics,
     the battery packs, or the sample
     pump (169 of the 232 samples lost
     due to monitor malfunction), these
     monitors will probably become ac-
     ceptable for future projects  of this
     type  providing that the recom-
     mendations discussed in the report
     are successfully  incorporated  into
     the monitor design. Excessive cali-
     bration drift accounted for the re-
     maining 63 of the 232 samples lost
     due to monitor malfunction (approxi-
     mately 6  percent of the  samples
     attempted. The monitors exhibited
     high  linearity (calibration r2  ss
     0.9997),  acceptable stability (86
     percent within ± 10 percent  of
     initial response  levels after  24
     hours), and reasonable  precision
     (median standard deviation of dupli-
     cate measurements = 0.25 ppm)
     during field monitoring.
 —  The GE/HP monitors will  probably
     be acceptable for such monitoring
     following  perfection of the  design
     and  incorporation of the recom-
     mendations suggested in the report.
     The  full  user-programmability  of
     these monitors will add desirable
     flexibility, not achievable  with the
     COED-1,  to future monitoring proj-
     ects. On-Board  micro-processor
     monitoring of, and compensation
     for, parameters such as cell temper-
     ature and battery voltage may in-
     crease monitor stability and preci-
     sion.

  Concerning the monitoring of alveolar
carbon monoxide by the method utilized
during  this project,  the following con-
clusions were reached:
 —  The  proposed  method  performed
     well, producing a mean difference
     between duplicate samples of 0.11
     ppm ±0.13 ppm  at the 95 percent
     confidence level and an estimated
     accuracy  of ± 0.3 ppm at  3.5 ppm
     and  ± 1.0 ppm  at 40 ppm.  The
     proposed modification to  the pro-
     cedure concerning use of humidi-
     fied zero  and calibration  matrices
     is, however, deemed necessary for
     procedural stability. The method is
     highly reliable (97.5  percent suc-
     cessful sample completion rate).
  Using the data collected in the Washing-  A
ton, D.C. and Denver metropolitan areas  "
with the Household Screening Question-
naire, weighted estimates were computed
of population characteristics. These esti-
mates were based  on screening inter-
views in 4394 households in Washington
and  2128 households in Denver.  In
particular, the population estimate for the
number of households in the two areas
was 953,71 4for Washington and 345,1 63
for Denver. Population estimates of per-
centages of  households with various
characteristics were as follows:

                   Washing-
                      ton     Denver

Use Fireplace          33%     30%
Use Wood Stove        4%      6%
Use Gas  Furnace       56%     71%
Use Gas  Stove         64%     25%
Use Gas  Hot Water     57%     78%
Have Attached         22%     35%
  Garage or
  Multi-Family
  Garage

  In addition to household characteristics,
several estimates were also obtained for
individuals' characteristics  in  the two  4
areas. For example,                     *

                   Washing-
                      ton     Denver

Male                 48%      47%
Smokers  (13 years     33%      38%
  or older)
Work (13 years        70%      72%
  or older)
Travel 3= 3            84%      82%
  times/week
  Regarding estimates of CO exposure
for the winter of 1 982-83 in Washington,
D.C., a data base was constructed from
the raw CO levels by activity data which
consisted of hourly CO values  on 712
respondents, activity patterns and cor-
responding CO levels on 705 respond-
ents, and CO breath measurements cor-
responding to the PEM CO data on 659
respondents. These data were  used to
obtain estimates of CO exposure for the
population of inference — the adult (18 to
70 years old), non-smokers in the urban-
ized portion  of the Washington,  D.C.
SMSA. The size of this population was
estimated to be 1 .22 million individuals.
The results presented below are weighted
estimates which apply to this population.

-------
  The weighted average maximum hourly
PEM CO level in Washington, D.C. was
6.74 ppm (this was computed as the
weighted average of the maximum hourly
CO value for each  individual in  the
sample). The average maximum 8-hour
CO level was 2.79 ppm. The percentage
of the population with maximum hourly
CO values over the 35 ppm CO standard
was estimated to be 1.28 percent while
the percentage with an 8-hour maximum
over the 9 ppm standard was 3.9 percent.
  Estimates  were  also made for  sub-
groups of the population. Persons in high-
exposure occupations (about 4.6% of the
total population) generally exhibited high-
er CO exposure levels: it was estimated
that about 24%  of this high-exposure
group had 1-hour CO exposures above
the 35-ppm standard and that about 28%
exceeded the 8-hour standard. It was also
shown that  CO levels were generally
higher for commuters, especially for
those with larger amounts of travel. For
example, 8% of the commuters indicating
16 or more hours of travel per week were
estimated to have maximum 8-hour CO
concentrations over 9 ppm, whereas
less than 1 % of the non-commuters were
estimated to have such levels.
  Breath CO levels (taken at the end of
the sampling periods, usually in  the
respondents'  homes) for the adult non-
smoking population in Washington aver-
aged 5.12 ppm.  Slightly  higher levels
were  observed for persons  with  high
occupational exposures and for persons
with large amounts of travel.
  By combining PEM data with data from
individuals' diaries, estimates of both CO
levels and time  durations for various
activities and personal environments
were  made.  In general,  these  results
were consistent with a priori expecta-
tions. For example,  the  activities  "in
parking garage or parking lot" and "travel,
transit" had  the  highest average  CO
concentrations (6.93 ppm and 4.51 ppm,
respectively) while "sleeping" had an
estimated CO concentration of only .85
ppm. Among  the environments consid-
ered, the three with the highest average
CO concentrations were "indoor parking
garage," "outdoor parking area," and "in
transit-car." The average levels for these
environments were 10.36,4.67, and 5.05
ppm, respectively.
  Variation from  duplicate hourly PEM
measurements under field conditions
were also analyzed. An analysis of vari-
ance of this data which considered per-
son-to-person, hour-to-hour,  and mea-
surement variation indicated that about 5
to 6 percent of total variation among the
hourly duplicate readings  was due to
deviations in measurements made by two
PEMs at the  same hour for the  same
person.
   T. D. Hartwell, C. A. Clayton, R. M. Mitchie, R.  W.  Whitmore, H. S. Zelon, S. M.
     Jones, and D. A. Whitehurst are with Research Triangle Institute, Research
     Triangle Park, NC 27709.
   Gerald G. Akland is the EPA Project Officer (see below).
   The complete report, entitled "Study of Carbon Monoxide Exposure of Residents
     of Washington, DC and Denver, Colorado," (Order No. PB 84-183 516; Cost:
     $20.50, subject to change) will be available only from:
          National Technical Information Service
          5285 Port Royal Road
          Springfield, V'A 22161
          Telephone: 703-487-4650
   The EPA Project Officer can be contacted at:
          Environmental Monitoring Systems Laboratory
          U.S. Environmental Protection Agency
          Research Triangle Park, NC 27711

-------
United States                       Center for Environmental Research
Environmental Protection              Information
Agency                            Cincinnati OH 45268
Official Business
Penalty for Private Use $300
                                      Hi>   00003^9
                                      U  b  hMVIK  PkUTtCTlON  AG£iMCY
                                      KEblUN  5  L1HKAKY
                                      250  S  UtAKtSURN  STHhfcT
                                      CHICAGO  IL

-------