vvEPA
                                  United States
                                  Environmental Protection
                                  Agency
                                   Environmental Sciences Research
                                   Laboratory
                                   Research Triangle Park NC 27711
                                  Research and Development
                                  EPA-600/S2-80-201  May 1981
Project  Summary
                                  Ambient Air Non-Methane
                                  Hydrocarbon  Monitor
                                  Darrell Burch
                                    A real-time monitor has been de-
                                  veloped for measuring non-methane
                                  hydrocarbons (NMHC) in ambient air.
                                  The monitor consists of two basic
                                  instruments, a methane monitor and a
                                  flame-ionization detector (FID). The
                                  methane monitor, which is based on
                                  gas-filter correlation techniques, makes
                                  use of the infrared absorption charac-
                                  teristics of methane to  measure its
                                  concentration. A slight interference in
                                  the measurement of methane by H2O
                                  vapor in the sample air is minimized
                                  using an electronic correction derived
                                  from a simultaneous measurement of
                                  the H2O concentration. The flame-ion
                                  ization detector measures the concen-
                                  tration of the total hydrocarbons
                                  (THC), including methane. The con-
                                  centration of non-methane hydrocar-
                                  bons is obtained by subtracting the
                                  methane concentration from the THC
                                  concentration. The noise-equivalent
                                  concentrations  (peak-to-peak) of the
                                  methane monitor and the FID are ap-
                                  proximately 50 ppb and 5 ppb of
                                  carbon, respectively. The estimated
                                  uncertainty in the measurement of a
                                  typical low-level NMHC concentration
                                  is between 20 ppb and 50 ppb. Con-
                                  centrations as high as 70 ppm can be
                                  measured.
                                    This Project Summary was devel-
                                  oped by EPA's Environmental Sci-
                                  ences Research Laboratory, Research
                                  Triangle Park, NC, to announce key
                                  findings of the research project that is
                                  fully documented in a separate report
                                  of the same title (see Project Report
                                  ordering information at back).
                                  Introduction

                                    Hydrocarbons in the atmosphere play
                                  an important role in the production of
                                  photochemical smog and are of great
                                  interest to atmospheric chemists and to
                                  those concerned with air quality. Meth-
                                  ane, the  most abundant atmospheric
                                  hydrocarbon, is essentially non-reactive
                                  at the normal ambient concentrations of
                                  a few ppm; thus, this gas does  not
                                  contribute significantly to atmospheric
                                  photochemistry. It  follows that  the
                                  quantity of most interest is the concen-
                                  tration of the other hydrocarbons, com-
                                  monly called non-methane hydrocarbons
                                  (NMHC).  For some detailed studies, it
                                  may be desirable to know the concen-
                                  tration of certain hydrocarbon species,
                                  but for many purposes, it is sufficient to
                                  know the sum of the concentrations of
                                  all the NMHC's.
                                    No convenient and reliable method
                                  has yet been developed for routine
                                  monitoring of NMHC's. One widely used
                                  instrument, the flame-ionization detector
                                  (FID), is capable of measuring the con-
                                  centrations of all the hydrocarbons,
                                  frequently called total hydrocarbons
                                  (THC). However, this quantity is not the
                                  one of most interest because it includes
                                  the inert methane, which may constitute
                                  from 30 to 90% of the THC.
                                   The Ford Aerospace and Communica-
                                  tions Corporation-Aeronutronic Division
                                  of Newport Beach, California, with  the
                                  support of the Atmospheric Chemistry
                                  and Physics Division of the Environ-
                                  mental Sciences Research Laboratory,
                                  Research  Triangle Park, North Carolina,

-------
has developed a system for measuring
the concentration of non-methane
hydrocarbons in an air sample. This
program involves coupling a commer-
cially available FID to measure the THC
concentration with a custom-designed
methane monitor. The NMHC concen-
tration is equal to the difference between
the concentrations measured by the two
instruments. The methane  monitor
uses gas-filter correlation techniques
and contains a set of multiple-pass
optics to give an 845 cm absorption
path. Water vapor in the air  sample
interfers with the methane measure-
ment giving a false reading that depends
on the partial pressure of the H20 vapor.
An infrared H2O monitor built as an
integral part of the methane monitor
measures the H2O concentration; this
measurement is used to correct the
apparent methane concentration. A
diaphragm pump circulates either sam-
ple air, bottled zero-gas or bottle span
gas through the monitors. The span
calibrations of both the methane monitor
and the FID are quite stable. Slight drifts
in the zero settings of both instruments
make it necessary to flush the sample
chambers with zero-gas approximately
once each  hour if the most accurate
results are required. The estimated
uncertainty in a measurement of a
typical NMHC concentration below 1
ppm (parts per million of carbon) is
between 20 and 50 ppb (parts per billion
of carbon).

Tests and Performance
  Results of a few of the tests performed
on the entire instrument with the meth-
ane monitor and FID  coupled together
are listed below. Output signals were
recorded separately for each monitor
and are expressed in terms of ppm of
methane. The sample cell of the methane
monitor was operated at 2-atm pressure
and 50°C.

Performance of Methane
Monitor Plus FID

Sensitivity and Noise
  (Peak-to-peak noise level with  3 sec
  electronic time constant)
  Methane
  Monitor:  0.05 ppm
  FID:      0.005 ppm

Linearity
  Methane Linear for concentrations
  Monitor:  less than 10 ppm, only a
           slight deviation from lin-
           earity for concentrations
           between 10 to 20 ppm.
  FID:      Not checked carefully, prob-
           ably linear to beyond 20 ppm.

Interferences
  Methane (Without automatic correc-
  Monitor:  tion) 2.5 percent H20 pro-
           duces interference corre-
           sponding to approximately
           +2.4 ppm. (With automatic
           correction) No interference
           for 1 percent H20. Less than
           +0.1 ppm for lower H2O con-
           centrations; approximately
           0.2 ppm at 3 percent H20.
  FID:      No significant interferences
           by normal atmospheric con-
           stituents.

  Additional topics covered in the main
report include: (1) the use of a Perma-
Pure Dryer as a means to reduce H20
interference in methane measurement;
(2) the modification of an FID to increase
stability; and (3) the electrical and
optical designs used in instrument
fabrication.

Conclusions
  An instrument consisting of a methane
monitor combined with a FID can be
designed and built with adequate sensi-
tivity and accuracy to monitor ambient
NMHC concentrations under most con-
ditions of interest. The low concentra-
tions of hydrocarbons make it necessary
that both of the instruments be very sen-
sitive and stable. It is desirable that the
instruments be kept in a temperature
controlled room and the sample air be
drawn in from the outside through a
heated line. Care must be exercised to
avoid contaminating  air samples and
calibration gases or losing hydrocarbons
on the walls of the gas handling system.
Activated charcoal filters in the fuel line
and  combustion-air line of the FID
remove  any residual hydrocarbons in
these gases and lead to greatly improved
instrument  stability. A single pump of
the proper design can circulate the air
sample through both the methane moni-
tor and  the FID. No changes  in the
hydrocarbon concentration of  an air
sample  appear to take place when it
passes through a diaphragm pump with
the interior Teflon  coated and properly
cleaned copper tubing heated to approx-
imately 50°C.
  A standard sensor (combustion cham-
ber plus electronics) for a FID provides
adequate sensitivity when operated
with a convenient  fuel mixture of 40%
H2 + 60% He. The short-term peak-to-
peak noise (period less than 10 sec) can
be made less than the equivalent of 5
ppb of methane for the FID and less than
50 ppb  of methane for  the methane
monitor. Longer-term drift of the zero-
settings of the  instruments normally
leads to uncertainties in  the measure-
ments that are  somewhat  larger  than
those imposed by noise unless the drift
is accounted for by flushing the sample
chambers once every few minutes with
zero-gas. Stability of the methane moni-
tor is improved greatly by controlling the
temperatures of the sample cell, band-
pass filter, and gas-filter  cell of the
methane monitor.
  Interference by H2O in the air limits
the accuracy of the methane  monitor
unless most of this gas is removed from
the air before it enters  the methane
sample cell. One acceptable method of
accomplishing this  is to pump the air
through a Perma-Pure Dryer before it
enters the methane sample cell. Air
going to the FID should by-pass the
dryer to avoid possible  adsorption  of
some of the complex hydrocarbons on
the walls of the dryer. The H20 inter-
ference  can also be  accounted for by
measuring the H20 concentration in the
methane sample cell and applying a
correction based on interference data
obtained previously with  samples of
H2O plus clean air.

Recommendations
  Additional tests should be carried out
with the instrument under a variety of
laboratory and field conditions to gain
more information about the detailed
performance. After these tests have
been completed, a prototype instrument
should be designed and built to operate
on the  same basic principles as the
present instrument. This  instrument
should include  a FID to  measure THC
concentrations and a methane monitor
that employs gas-filter correlation tech-
niques.
  The following features and procedures
are recommended  for the prototype
instrument. Many of these features are
included in the present instrument and
have proven to be desirable; others are
recommended as a result of knowledge
gained while assembling and testing
the present instrument. Important fea-
tures of the present instrument that are
not mentioned below should be included.
1.  Package both the FID and methan^
   monitor into a single unit.

-------
 2. Use the combustion chamber from a
    commercially available FID with
    combustion fuel of 40% H2 and 60%
    He.
 3. Pump sample air through heated
    lines to both the methane monitor
    and the FID with a single diaphragm
    pump capable of producing pressures
    up to 5 atm. Split the gas flow so that
    the  gas  to the FID does  not pass
    through  the methane sample cell.
    Include in the line to the FID a small
    "delay tank" so that at a given time
    the FID is  sampling air that entered
    the inlet line over the same period of
    time as the air in the methane sample
    cell.
 4. Pass the air to the methane sample
    cell through a dryer such as a Perma-
    Pure Dryer to remove most of the
    H20 vapor and thus reduce the inter-
    ference by this gas in the measure-
    ment of methane concentration.
 5. Employ activated charcoal  filters, or
    some substitute, in the lines for the
    combustion-air and fuel.
 6. Operate the sample cell of the meth-
    ane monitor between 3 atm and 5
    atm to increase sensitivity and reduce
    interference due to residual H20.
    This also improves the efficiency of a
    dryer similar to the Perma-Pure
    Dryer.
 7. Decrease the volume of the methane
    sample cell and use multiple-pass
    optics in the cell to obtain  a sample
    path length between approximately
    8m and 15m.
 8. Shape the  sample cell to reduce the
    volume while passing approximately
    the same amount of radiation in the
    monitoring beam as the present
    instrument.
 9. Control  the temperatures of: (a)
    methane gas-filter cell, (b) spectral
    bandpass filter for methane monitor,
    (c) methane sample cell, and (d)
    regulators, valves, tubing,  etc., that
    are parts of the FID. Heat gas lines,
    including the dryer, to approximately
    50°C; it is not necessary to control
    the temperature  of these lines.
10. Include electronics to measure di-
    rectly the  difference between  the
    output signals of the FID and methane
    monitor;  this voltage is proportional
    to the NMHC concentration.
Da rre I I Burch is with Ford Aerospace & Communications Corporation Newport
  Beach, CA 92660.
William McClenny is the EPA Project Officer (see below).
The complete report, entitled "Ambient Air Non-Methane Hydrocarbon Monitor,"
  (Order No. PB 81 120-008; Cost: $6.50, subject to change) will be available
  only from:
        National Technical Information Service
        5285 Port Royal Road
        Springfield. VA 22161
        Telephone: 703-487-4650
The EPA Project Officer can be contacted at:
        Environmental Sciences Research Laboratory
        U.S. Environmental Protection Agency
        Research Triangle Park, NC 27711
       ; US GOVERNMENT PRINTING OFFICE. 1W1 -757-012/7121

-------
United States                     Center for Environmental Research
Environmental Protection             Information
Agency                          Cincinnati OH 45268
                                                                                      Agency
                                                                                      EPA 335
Official Business
Penalty for Private Use $300
         PS   0000329
         U  S  t,NVIR  PROTECTION
         REGION  5 LIBRARY
         230  S DEARBORN  STREET
         CHICAGO It, 606U4

-------