United States
                 Environmental Protection
                 Agency	
Atmospheric Research and
Exposure Assessment Laboratory
Research Triangle Park NC 27711
                 Research and Development
EPA/600/S3-89/016Aug. 1989
&ERA        Project Summary
                 Analytical Method
                 Evaluation for Measuring
                 Ethylene Oxide  Emissions from
                 Commercial  Dilute-Acid
                 Hydrolytic Control Units
                 Joette Steger
                  The Source Branch of the U.S.
                 Environmental Protection Agency's
                 (EPA) Environmental Monitoring
                 Systems  Laboratory (EMSL)  at
                 Research  Triangle Park,  North
                 Carolina, has a program to develop
                 stationary source test methods of
                 known precision and accuracy for
                 determining compliance  with EPA
                 standards. While participating  in this
                 program,  Radian  Corporation
                 performed  a field  evaluation of a
                 method for sampling and analyzing
                 ethylene oxide (EO) In the vent
                 stream from a dilute acid scrubber of
                 a commercial sterilizer. The purposes
                 of the test were to evaluate the ability
                 of the analytical  method  for
                 quantitatlng EO in the  scrubber
                 emissions, to determine EO stability
                 in sample containers, and to measure
                 the EO concentration profile and EO-
                 to-oxygen    and    EO-to-
                 dlchlorodifluoromethane (CFC-12)
                 ratios during  the  initial sterilizer
                 evacuation.
                  The test  method  acquired  three-
                 second grab  samples  that were
                 analyzed by gas chromatography with
                 flame lonization detection (GC/FID).
                 Samples were acquired at  one-,
                 three-, four- or six-minute intervals.
                 All samples were analyzed in the field
                 and a portion  of the samples were
                 reanalyzed in the laboratory.
  The analytical  method  was
evaluated on three 1/8-inch  (32
millimeters)  OD stainless steel
columns packed with a liquid phase
coated on 60/80 mesh Carbopack B.
The 5% Fluorcol column was 10 feet
(3 meters) long;  the  3% SP-1000
column was 8 feet long (2.4 m); and
the 1% SP-1000 column was 6 feet
(1.8 m) long. Evaluations were based
on column efficiency, resolution,
linearity, retention  time stability, and
limits of detection and quantitatlon.
Sample stability was evaluated  in
one-liter Tedlar bags, 5-mL gas-tight
syringes, and Vacu-Samplers. The EO
concentration  profile was measured
by taking  samples  at one-minute
intervals during Test 1 Evacuation 1.
This test used 12 wt% EO and 88 wt%
CFC-12 as the sterilizing gas.
  The full  report  contains
conclusions and recommendations
based on  the field  test  results, a
description of the field test site, the
grap sampling and analytical method
used, results of the field evaluation,
test data,  and references used  to
prepare the report
  This Project  Summary  was
developed by EPA's Atmospheric
Research  and Exposure Assessment
Laboratory, Research Triangle Park,
NC, to announce key findings of the
research  project  that  is fully
documented in a separate  report of

-------
the same title (see  Project Report
ordering information at back).

Introduction
  Because  EPA has  listed  EO  as a
possible hazardous  air pollutant, a
standardized  analytical  method  is
needed to measure EO levels in a vent
stream  and for determining  control
equipment  performance. The  method
tested acquired  three-second  grab
samples that were analyzed  by  GC/FID.
The purposes of the field evaluation were
to evaluate the analytical method  using
samples from an operating  commercial
sterilizer with a  dilute acid scrubber, to
determine  EO  stability  in  sample
containers as a function of storage time,
and to measure the  EO concentration
profile and  the  EO-to-oxygen  and
EO-to-CFC-12 ratios  during  the  initial
sterilizer evacuation.

Procedures

Facility Description
  The field evaluation was conducted at
Burron  Medical,  a  medical  supply
sterilization facility, located in Allentown,
Pennsylvania. The facility  has  three
1000-ft3 (28-m3) sterilizers,  one  uses
a 12/88 (wt%/wt%) EO and CFC-12 gas
mixture and two use 100% EO.
  The exhaust  from  the sterilizers  is
controlled by a  DEOXX™" system. The
DEOXX system is a dilute acid scrubber
manufactured  by  Chemrox,  that
hydrolyzes the EO to ethylene glycol. At
the time  of  the  test the  scrubber
contained dilute sulfuric  acid  and
approximately 10 wt% of ethylene glycol.
  Each chamber is equipped with a total
recirculating  oil-sealed pump.  The
pumps are equipped with  gas/liquid
separators that emit  the gas to the
DEOXX system and recirculate the liquid
to the pump inlet.
  The two different sterilization programs
used for testing  are shown in Figures 1
and 2. Three tests were performed, one
with product in the chamber using  100%
EO and two without product using  12/88
gas. Data from  all of  these  tests were
reduced and used to prepare this report.

Sampling Location
  Samples  were  acquired  after the
scrubber, at a point as close to the end
of the plant's stack as was feasible. The
stack  was  modified  by  installing
 "Mention of trade names or commercial products
 does not constitute  endorsement or  recom-
 mendation for use
additional pipe to remove exhaust gases
from  the  sampling areas and  orifice
plates to provide a small amount of back
pressure  to  facilitate sampling and
eliminate the  possibility of diluting the
sample with ambient air.

Sampling Procedures
  Grab  sampling  was  used  for
determining EO,  CFC-12, and oxygen
concentrations. For EO  and CFC-12
measurements, sample was  withdrawn
into  a  heated,  1/4-inch   (64-mm),
Teflon®  line  using  a  Teflon-lined
diaphragm pump. A stainless steel, 1/4-
mch  (64-mm)  tee was used  after the
pump to provide  a continuous purge of
the vent  gas through the  sampling
system. After the tee, on one branch, the
sample stream  was  routed through
Tygon  tubing  and exhausted downwind
from  the sampling area. There  was a
pinch  clamp on  the  Tygon  tubing  to
block  the  exhaust  line during sample
collection.  On  the other branch  of the
tee,  a stainless  steel,  1/4-inch  (64-
mm),  toggle-operated shut-off  valve
was placed prior to the  sample container.
  The  EO and CFC-12 sampling line
was  continually  flushed  with  sample
throughout the test. Prior to attaching the
sample container,  the  line  was flushed
with sample by closing the pinch clamp
on the Tygon exhaust  line and opening
the toggle  valve on the sample line. The
same  procedure  was  used  to fill  the
sample container.
  BACHARACH Fyrite  oxygen indicator
was used to determine percent levels of
oxygen in the sterilizer exhaust at the
sampling  location.  The  oxygen  was
chemically absorbed from the  sample by
chromous chloride. Accuracy of analysis
was  ± 0.5 percent. Percent  levels  of
oxygen were measured every time a bag
sample was taken to indicate the dead
volume of the  scrubber system  and to
determine  EO-to-oxygen ratios.  An
aspirator bulb  removed sample from the
stack  downstream of the  EO sampling
port.
  Exit gas temperatures were measured
using a bimetallic  temperature probe and
a pyrometer.  Stack  temperatures were
digitized by a  calibrated pyrometer and
recorded every time a bag sample was
taken.

Analytical Procedures
  The  analytical  method  used  for
measuring  EO and CFC-12  was gas
chromatography  with  flame  lonization
detection (GC/FID). A  dual FID  Varian
3400 GC was equipped with  a 10-ft (3-
m) by 1/8-inch (32-mm)  OD stainless
steel column containing 5% Fluorcol (
60/80 Carbopack B  and a 6-ft  (1.8-n
by  1/8-inch  (32-mm)  OD  stainles
steel column containing 1% SP-1000 c
60/80 mesh Carbopack B. On  July 9, tr
1% SP-1000 column  was replaced wi
an  8-ft  (2.4-m)  by   1/8-inch (32-mn
OD stainless steel column containing 3'
SP-1000  on 60/80 mesh Carbopack I
The FID electrometers were connected I
Shimadzu CRI-A  integrators.
  The GC  column  oven was operate
isothermally at 45°,  55° and 65QC  c
June 21,  22  and  23, respective!'
Laboratory  reanalyses were  performe
with an isothermal column temperature i
65°C except on  July  9 when  the  3'
SP-1000 column was  used at 55°C.  Tr
injector oven was heated to  I75°C,  ar
the detector oven to  200°C. Nitroge
carrier gas flow rates  were 30 mL/mi
The  FID  support  gas  flow  rate
recommended  by the  GC manufactun
were used. During Test  1 Evacuation
when both EO  and  CFC-12 wer
quantitated, the FID  electrometers wei
programmed so  that the range  woul
shift from 10'10  or  10'11  amperes 1
10~9 amperes.  One-milliliter sample
were  injected  using  a 5-mL gas-tigl
syringe with an on/off valve.

Gas Chromatograph Calibrator
  Both channels  of  the  chromatograp
were calibrated for EO at the beginnin
and end of the day. On the first test da'
the GC/FID was also calibrated for CFC
12  to determine the  EO-to-CFC-1
ratios. At  least one  standard was als
analyzed  after  every  twentieth  sampl(
Calibration curves  consisted of
minimum  of  three  standards  the
bracketed the sample concentrations.

Test Results and Conclusions
Linearity
  Based on a one mL injection volum<
the  FID  response using these thre
columns was linear between  31.25  an
100  vol%  for CFC-12. Below  31.2
vol%  CFC-12  linearity  was  nc
evaluated. Using  a one-milliliter injectio
volume, the FID response when using th
3%  SP-1000 column had the  large:
linear range (0.5 to 100 vol% and 2.28
to 1,254  ppmv)  for  EO  at the  percei
level and  at the part-per-million  (pprr
level. At  the  part-per-thousand (pp
level, the  FID  response using  the  1°
SP-1000  column  had the largest lines
range (22.81  to  5.016 ppmv)  for  EC
Therefore, the  FID response using  th
1% SP-1000 column provides the be:
linear range  for  determining  th

-------
                      5
                      4)
                      I
                                                  Chamber Program
                                                Test Two— 12/88 Gas
                                    40
                                            80
                    Figure  1.
                    120     160    200

                    Elapsed time (minutes)

Chamber pressure during Test 2 using 12/88.
                                                                           240
   280
                                                                                          320
maximum EO concentration emitted from
a dilute-acid hydrolytic control unit.

Efficiency
  Of the three  columns evaluated, the
5% Fluorcol  column  operated  at 65°C
provided the best efficiency for analyzing
percent levels of EO.  For ppm  levels of
EO the two columns  were comparable.
The  3% SP-1000 column was clearly
the least efficient column  at all tested
temperatures  and  concentrations.
Therefore, the Fluorcol column  provides
the best efficiency for percent level EO
analyses and the 1% SP-1000 column
provides the best efficiency for ppm level
EO analyses.

Resolution
  The   1%  SP-1000  and  the   5%
Fluorcol columns provided comparable
resolution of the EO from the CFC-12 at
the percent and ppt  levels  At column
temperatures  of 45° and 55°C  baseline
resolution of EO from  CFC-12  was
obtained on both columns Both of these
columns provided  significantly better
resolution of the  two compounds than did
he 3%  SP-1000 column. Therefore, the
J%   SP-1000  column   is   not
        recommended for analyzing samples of
        12/88 gas. Resolution was not evaluated
        at the ppm level.

        Retention Time Stability
           The EO  retention  times  were  most
        stable on  the  1%  SP-1000 column
        where they  varied from  1.38 to 1.46
        minutes (± 6%) at 55°C and were least
        stable on the  5 Fluorcol column  where
        they varied  from 2.21 to 5.43 minutes (±
        150%)   at  55°C. In  general the EO
        retention times varied least  at column
        temperatures  of 65°C   The  CFC-12
        retention times  were most stable on the
        3% SP-1000  column  where they  varied
        from  1.46 to 1.56 minutes  (± 7%) at
        65°C and  were least stable on the 5%
        Fluorcol column where they varied from
        4.72 to 557 minutes (±  18%) at 45°C.
        Therefore, the  Fluorcol column  where
        they varied  from 4 72 to 5 57 minutes (±
        18%) at  45°C. Therefore,  the Fluorcol
        column should only be used to analyze
        for EO at the  percent levels  in situations
        where the concentration is not expected
        to change by  more than two orders of
        magnitude.
Limits of Detection and
Quantitation
  The  3%  SP-1000  column  provided
the lowest detection and quantitation
limits, therefore,  it would be  the best
column for low-level and  trace  analyses.
The limits of  detection and  quantitation
were comparable  for  the 1%  SP-1000
and 5%  Fluorcol  columns;  therefore,
these  two   columns   may  not  be
distinguished by these criteria.

Ethylene Oxide Stability
  Evaluations  were made of the stability
of EO in samples collected in  1-L Tedlar
bags,   5-mL  gas-tight  syringes, and
Vacu-Samplers.  After  4 days  in an
unpunctured  Vacu-Sampler four  out  of
seven  duplicate samples or 57% were
within  ± 8% of the EO concentration in
the original samples. The seven duplicate
samples  in   the  unpunctured  Vacu-
Samplers lost an average of 15%  of the
EO.  Therefore, EO  samples  in
unpunctured  Vacu-Samplers  remained
stable for  up to  4 days, with stability
defined as over  50% of the  samples
being  within  ±  10% of the original
concentrations and the average of all EO

-------
                                                     Chamber Program
                                                   Test Three — 100% EO
                      Figure 2.
                     200
                  Elapsed time (minutes)

Chamber pressure during Test 3 using 100% EO.
                                                                               400
concentration changes being within  ± 15
percent.
  After 4  or 5  days m  gas-tight
syringes, only 2  out  of 12  duplicate
samples or 17% were within  ± 9% of
the EO concentration  in  the original
samples. The samples  in  the  syringes
lost an average  of 20%  of the  EO.
Therefore, samples are not  stable in
syringes for 4 or 5  days.
  Stability results  are listed in Tables 1
and  2  for bag  samples  containing
exhausts from 12/88 and 100% EO tests,
respectively. In  leak-free bags the EO
in  12/88 exhaust  at EO concentrations
above  1,000 ppmv   varied by  an
averaged relative  difference  of  +  9%
after 48 hours. After 5 to 6 days the bags
with concentrations of  EO  greater than
1,000  ppmv showed an average EO loss
of  22  percent. In the same time period,
bags with concentrations of  EO less than
500 ppmv showed  an average EO loss of
36 percent. After 17 to 18 days the bags
with  EO concentrations greater  than
1,000  ppmv showed an average EO loss
of  41   percent.  Therefore,  samples of
12/88 exhaust  containing EO  at
concentrations  above 1,000  ppmv are
stable in Tedlar bags for up  to 48 hours.
         Leak-free bags containing  100% EO
       exhaust  at EO  concentrations above
       1,000  ppmv  varied  by  an averaged
       relative difference of  + 3% after  four
       days. After 8  days these bags averaged
       a 20% loss and after 12 days an average
       of 33% of the EO was lost. In  leak-free
       bags  of  100%  EO  exhaust  at
       concentrations  below 500 ppmv  an
       average of approximately 20% of the EO
       was lost after 4 days. No additional loss
       was observed in these bags  between
       Day 4 and 12.  Therefore,  samples of
       100%  EO  exhaust containing EO at
       concentrations above  1,000  ppmv are
       stable for at least 4 days.
         Since  the  EO loss  in  the  bags is
       concentration  dependent and more EO is
       lost at lower concentrations, the EO loss
       is expected to be caused by adsorption
       of the EO onto the bag  walls. Initially, an
       equilibrium probably occurs between the
       bag surface and  the gas within the bag
       accounting  for the 2-  to  4-day period
       of stability observed As the gas remains
       m  the bag longer,  the  EO  may  be
       penetrating further  into the bag  wall,
       accounting  for the additional  EO  loss
       The sample stability was  decreased by
the  high  surface-to-volume ratio  m th
1-L  bags.

Ethylene Oxide Concentration
Profile
  During the first evacuation  of the fir
test  using   12/88  gas,   the  E
concentration ranged from 400 ppmv  I
1,500 ppmv and reached  a  maximui
plateau  between  13 and  20  minute
Concentrations  of  EO during   I2/8
exhausts were approximately 1,200 ppm
during  the  second  exhaust,  600  pprr
during  the third  exhaust,  300  ppm
during  the fourth exhaust,  100  pprr
during  the  fifth exhaust, and 80  pprr
during the sixth exhaust. During the fir
evacuation of the third test using  100'
EO,  the EO concentration  ranged froi
500 ppmv to 3,000 ppmv Concentratior
of EO  during  100% EO exhausts wei
approximately  600  ppmv  during th
second exhaust,  and 300  ppmv  dunr
the  third exhaust. Therefore, to measui
the  maximum EO concentration emitte
samples should be acquired between  1
and 20 minutes after the start of the fir
chamber evacuation when 12'88 is used
  Ratios  of  the  EO-to-CFC-1
concentration were calculated  in  12/f

-------
exhaust and a plot of the ratios versus
time  was  similar to  a plot of  EO
concentration versus time due to the high
and   relatively  constant  CFC-12
concentration. Oxygen concentrations in
the 12/88 exhaust were  monitored using
Fyrite  and  indicated  that after five
minutes the exhaust contained minimal
air. Therefore,  EO-to-CFC-12  ratios
could  be established for determining the
maximum EO concentration emitted from
chambers  using  12/88. The EO-to-
CFC-12 ratios would  only  be useful
during the first  evacuation when  the
expected CFC-12 concentration is 99.9
percent.

Recommendations
  Based  on the  test  results, Radian
makes the following 7 recommendations.
  1 A 5%  Fluorcol column  at  65°C is
   recommended  for measuring per-
   cent level EO and CFC-12 concen-
   trations,  levels expected at  the
   scrubber inlet or  m  the exhaust of
   uncontrolled  sterilizers.  The FID
   response when using a 5% Fluorcol
   column is linear, as indicated  by an
   F-test  for linearity from 0.5  to 50
   vol% EO. The 5% Fluorcol column is
   the most efficient column tested for
   percent levels of  EO, and provides
   baseline resolution  of the  CFC-12
   and EO at column  temperatures of
   45 and 55°C  and  near  baseline
   resolution  at 65°C.  Also, the
   retention  time instability  on the
   Fluorcol  is  less  pronounced  at
   percent concentrations.
  2 A 1% SP-1000 column  at  45°C is
   recommended for  measuring ppm
   levels of EO,  levels  expected  at the
   scrubber outlet. The FID  response
   when using a 1%  SP-1000 column
   is linear from 22.81 to 5,016  ppmv
   EO. The  1% SP-1000  column  is
   the most efficient  column tested for
   ppm  levels  of EO, and  provides
   baseline resolution  of the  CFC-12
   and EO at 450C.
  3. It  is recommended  that  a  3% SP-
   1000 column at 45°C be investigated
   for quantitating sub-ppm levels  of
   EO, levels expected  in ambient air at
   sterilization facilities. The  3% SP-
   1000 column  provided  limits of
   detection and  quantitation almost
   three times lower than the other two
   tested columns.
  4 It  is recommended  that  grab
   samples  collected m  gas-tight
   syringes be analyzed within several
   hours of collection, samples col-
   lected in bags  within 48 hours of
   collection and samples collected in
    Vacu-Samplers within  4 days of
    collection.  Syringe samples  lost an
    average of 20%  of the EO  after 4
    days.  In  leak-free  bags,  EO
    remained stable for at least 48 hours
    and may be stable for up to  4 days.
    After 5 or  6 days, bag samples lost
    an average of  one-third of the  EO.
    In unpunctured  Vacu-Samplers,
    57% of the samples were within ±
    8% of the expected  concentration
    after 4 days and the  average EO
    loss for all samples was 15 percent.
  5. Sampling  at  least  three  times
    between 13 and 20 minutes after the
    first indication  of  flow through the
    stack  is recommended   for
    determining   maximum   EO
    concentration emitted during  the first
    evacuation. The EO concentration in
    the exhaust   increased linearly
    between  5 and  13 minutes,   pla-
    teaued between 13 and 20 minutes,
    and dropped off  after 20 minutes.
    This time  frame  corresponds to
    chamber pressures between  16 and
    10 pounds per square  inch actual
    and may need  to be adjusted when
    using  a  different  sterilization
    program.
  6. If  the  second  and  following
    evacuations are monitored, sampling
    between 5 and 10 minutes  during
    these  evacuations  is recommended.
    During the  first 5  minutes, the  gas
    coming out of the stack  is a  mixture
    of gas  from   the chamber   and
    residual gas from the  scrubber so
    samples taken  during this time may
    be biased high or low.
  7 To  determine   EO-to-CFC-12
    ratios  during the  first evacuation,  a
    0.1-mL sample volume is   recom-
    mended for CFC-12 quantitation. A
    1-mL  sample of  percent levels of
    CFC-12 overloads the  analytical
    column, decreasing the column  per-
    formance.
  The full report was submitted in partial
fulfillment  of EPA   Contract  No.  68-02-
4119 by  Radian Corporation under the
sponsorship of  the  U.S.  Environmental
Protection Agency. The report covers the
period from March to August of 1988.

-------
Table 1.
Sample
Bag No.
13
20
22
25
30
32
36
37
39
41
44
45
48
50
55
56
57
Ethylene
June 21
1230
1340
1200
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
NA
Oxide Stability in Tedlar Bags Containing 12/88
Concentration (ppmv)
June 22
1370
NA
1360
976
374
133
28.5
36.7
29.8
1170
NA
1240
NA
305
NA
NA
NA
June 23
1320
NA
1320
NA
NA
96.5
NA
NA
27.2
NA
NA
1380
NA
340
80.6
101
83.5
June 27
995
1010
956
734
NA
34.2
NA
NA
1.08
NA
1120
958
462
214
57.4
53.6
51.8
July 1
824
830
828
562
NA
34.4
61.8
NA
11.9
636
954
865
370
195
75.2
70.5
65.8
Exhaust
Julys
716
1420
790
962
166
NA
48.8
124
NA
546
803
690
556
173
71.0
70.6
48.4

July 9s
733
985
760
536
NA
NA
65.6
NA
NA
538
810
683
NA
NA
NA
NA
NA
a Samples analyzed on a 3% SP-1000 column
NA  = Not available

-------
Table 2. Ethylene Oxide Stability in Tedlar Bags Containing 100% EO Exhaust
Concentration (ppmv)
Sample
Bag No.
58
60
63
64b
65
66
67
68
69
70
71
72
June 23
343
2,840
544
342
344
334
430
426
387
1,530
1,440
1.200
June 27
NA
2,410
470
170
238
230
370
352
282
1,760
1,490
1.270
July 1
270
2,030
426
13.7
258
268
322
307
270
1,400
1,080
974
JulyS
248
1,660
430
NA
242
244
303
516
226
1.200
902
756
July 9s
2128
1,408.0
364.8
NA
220.8
225.6
2840
NA
NA
NA
NA
NA
a Samples analyzed on a 3% SP-1000 column
b Bag leaked
NA  = Not available

-------
  Joette Steger is with Radian Corporation, Research Triangle Park, NC 27709
  John H. Margeson is the EPA Project Officer (see below).
  The complete  report, entitled  "Analytical Method  Evaluation  for  Measuring
        Ethylene  Oxide  Emissions from Commercial Dilute-Acid Hydrolytic Control
        Units," (Order No. PB 89-155 253/AS; Cost: $21.95, subject to change) will
        be available only from:
            National Technical Information Service
            5285 Port Royal Road
            Springfield, VA 22161
            Telephone: 703-487-4650
  The EPA Project Officer can be contacted at:
            Atmospheric Research and Exposure Assessment Laboratory,
            U.S. Environmental Protection Agency
            Research Triangle Park, NC 27711
United States
Environmental Protection
Agency
Center for Environmental Research
Information
Cincinnati OH 45268
Official Business
Penalty for Private Use $300

EPA/600/S3-89/016
                                                                                                 .,_,

-------