United States
Environmental Protection
Agency
Industrial Environmental Research
Laboratory
Research Triangle Park NC 27711
Research and Development
EPA-600/S7-82-063   Feb. 1983
Project Summary
Evaluation  of  Potential  VOC
Screening  Instruments
K. T. Menzies and R. E. Fasano
  This report describes the evaluation
of potential fugitive source  emission
screening instruments for analysis of
volatile organic compounds (VOCs). An
initial review of available portable VOC
detection instruments  indicated that
detectors operating on several princi-
ples (i.e., flame  ionization,  catalytic
combustion, photoionization, infrared
absorption, and thermal conductivity)
might be  useful for VOC  analysis.
However, flame ionization and catalytic
combustion devices evaluated previ-
ously showed poor sensitivity for highly
substituted aliphatic  and aromatic
organic compounds. Instruments util-
izing photoionization and infrared may
be able to meet necessary criteria for
practical and accurate VOC analysis of
highly substituted organics. Therefore,
three commercially available instru-
ments (HNU  PI-101, AID 580, and
Foxboro Miran 80) were modified and
evaluated for 32 such  compounds at
concentrations of 100-10,000 ppmv.
Results  indicate  that photoionization
may be suitable for general VOC screen-
ing, but a reliable instrument/dilution
system does not exist. Infrared absorp-
tion will apparently not provide suitable
general  VOC  screening,  but may be
useful for analysis of some classes of
organic compounds.
  This Project Summary was developed
by  EPA's Industrial Environmental
Research Laboratory. Research Triangle
Park. NC, to announce key findings of
the research project that is fully docu-
mented in a separate report of the same
title (see  Project  Report  ordering
information at back).

Introduction
  The U.S. EPA has issued performance
standards and guidelines to limit emis-
sions of volatile organic compounds
(VOCs) from  several stationary  source
categories; e.g., surface coating  opera-
tions. It is  apparent that sources other
than classical point sources may also
emit large  amounts of  VOCs into the
workplace and surrounding atmosphere.
As described  in EPA Method 21, Deter-
mination of Volatile Organic Comppund
Leaks, technically and economically feas-
ible devices suitable for monitoring such
fugitive sources as valves include only a
few that are portable. These devices can
be placed near possible  points of emis-
sions and will respond to releases of the
organic compounds. Instruments suit-
able for this purpose include, but  are not
limited to, catalytic oxidation, flame
ionization, infrared absorption, and photo-
ionization detectors.
  Unfortunately, due to the chemical
complexity  of  many fugitive sources and
the lack of  universal sensitivity of these
detectors, the detectors previously eval-
uated cannot  adequately measure the
concentration of all chemicals likely to be
released. This fact has been documented
for two commercially available detectors
using flame  ionization  and catalytic
combustion.  Among  168 compounds
tested, 23 showed sufficiently poor
response that the actual and measured
concentrations differed  by a factor of
greater than five.
  The classes of compounds which show
the poorest agreement with the actual
concentration  generally incorporate func-
tional groups  such as halides, hydroxyl
(alcohols), carbonyl (aldehydes, ketones)
and carboxylate (acid) and include both
substituted aromatic hydrocarbons and
low-molecular-weight, highly substituted
aliphatic compounds.
  Additional  portable devices  which
respond accurately to these compounds

-------
are needed for VOC screening. Instru-
ments  other  than flame ionization or
catalytic oxidation detectors which might
meet this goal operate on the principles of
infrared absorption, photoionization, and
thermal conductivity.
  The first step in this task was to procure
one or more units of detectors that meet
the specifications of Method 21. The VOC
instrument must be rugged, reliable,
relatively inexpensive, portable, and easy
to operate. Of course, it must respond to
the organic compounds of interest and be
able to measure the leak definition concen-
tration  specified in the regulations.
According to Method 21, the instrument
must be intrinsically safe for operation in
explosive atmospheres as defined by the
applicable National  Electric Code.  Few
detectors are now "approved" for such an
environment.
  The second step in this task was to set
up a laboratory system capable of mixing
known volumes  of vapors with air and
delivering the mixtures of known concen-
tration to the detectors. Tedlar bags and a
volumetric mixing system were  selected
for sample preparation since they provide
adequate accuracy/precision and require
little cost or time to set up.
  The third step in this task was evalua-
tion of the detectors for response to the
compounds of interest. The response
factors (RFs) were determined at several
concentrations from 100 to 10,000 ppmv:

  Rp_     Actual Concentration
      Concentration Calculated from
           Instrument Response

 Measurements were limited to concen-
 trations approaching about 90% of the
 saturation  concentration of 75% of the
 lower  explosive  limit (LEL). To  permit
 statistically valid interpretation of the
 measured response factors, five replicate
 measurements at three concentrations
 were conducted. Data analysis included
 calculations  of  mean response  factors
 and confidence intervals.

 Results and Discussion

Photoionization Detection
   The photoionization  technique was
 evaluated for a  limited number of com-
 pounds due to both chemical and, more
 significantly,  equipment problems. The
 PI-101  was  calibrated with dichloro-
 methane  to  permit direct comparison
 with response factors previously report-
 ed. The response factors observed for the '
 16 compounds tested on the  PI-101
range from 0.50 to 48. Those for the 13
compounds evaluated at  1000 ppmv,
summarized in Table 1, range from 0.5 to
8.0. Of the compounds tested, 75% (12)
have response factors of less than 5.0
and greater than 0.2. There appears to be
no obvious trend of response factor with
molecular weight (carbon  number) or
functionality within this group. It is
interesting to note that, for both alcohols
tested (methanol  and ethanol), the re-
sponse factors are inversely proportional
to carbon number. This is consistent with
the large ionization potential and fewer
ionizable electrons in methanol, as
compared to ethanol. It also appears that
non-bonding electrons  on  the oxygen
atom  of the alcohols do not provide a
much greater photoionization yield than
other sigma-bonded  electrons  in  com-
pounds with similar carbon numbers.
  Although the  response factors for the
limited number  of compounds tested do
not unequivocally confirm the suitability
of photoionization as a general VOC
screening technique, an important but
cautious observation can be made. That
is, based on this small sample of com-
pounds tested, which includes an aro-
matic  compound  (toluene),  an  ether
(acetal), an alcohol (ethanol)  and chlo-
rinated alkanesftrichloroethane and chlo-
roform), the response factor at concen-
trations of  500 -  10,000 ppmv may be
within a factor of five.  This result  is
consistent with an expectation of more
similar photoionization yield from sigma
and pi electrons when the compound is
influenced  by UV  radiation of approx-
imately 12eV rather than 10eV. The
expectation that photoionization yield for
aliphatic and aromatic compounds may
be similar  indicates the potential use-
fulness of photoionization in VOC screen-
ing.
  In terms  of  suitability  as a potential
VOC detector, the most significant result
with  respect  to  the photoionization
detector (HNU  Systems,  Inc. PI-101 and
AID, Inc. 580) is the difficulty observed in
operating the prototype dilution system.
Both dilution probes were designed and
Table 1.   Response Factors with 95% Confidence Intervals*
OCPDB*
ID No.
-•
790
810
930
-
-
1660
1236
2500
-•
3349
3291
3395
Compound Name Response Factor
Acetal
Carbon Disulfide
Carbon Tetrachloride
Chloroform
Diketene
Dimethylsulfide
Ethanol
Ethylene Dichloride. trans 1.2
Methanol
Pentanethiol, 1-
Toluene
Tetrachloroethane, 1, 1,2,2-
Trichloroethane, 1,1,1 -
1.1
0.50
0.94
1.3
6.8
0.85
2.8
0.96
8.0
0.79
0.85
1.4
0.98
Confidence Interval
1.0 •
0.45-
0.77-
1.3 •
5.9 -
0.80-
2.4 -
0.84-
6.3 •
0.68-
0.67-
1.3 -
0.74-
1.1
0.57
1.20
1.4
7.9
0.90
3.4
1.1
11.0
0.96
1.2
1.4
1.4
* Concentration = 7000 ppmv; a/I are of the light liquid (LLj volatility class.
b Organic Chemical Producers Data Base.

-------
fabricated  by the respective manufac-
turers under severe time limitations.
Neither probe was designed to permit
reliable independent measurement of
dilution ratio or reproducible adjustment.
Thus, the  absolute dilution ratio  is in
some doubt. The ability to adjust the
dilution ratios was practically nonexis-
tent.

Infrared Detection
  A total of 32 compounds were analyzed
on  the  Miran 80. Prior  to testing, the
instrument was calibrated with individual
span gases  at eight analytical wave-
lengths which correspond  to individual
functional groups; e.g., C-H; C-C1; C-OH.
Test compounds were then run, and the
instrument response  calculated on the
basis of the  response indicated by the
specific span gas used at individual
analytical wavelengths.
  An analysis of the data indicates that
the response factors for most compounds
with a particular functional  group,
determined at an analytical wavelength
which corresponds  to that functional
group, are generally less than 20. This is
consistent  with the general observation
that the  functional  group  is  more
important  than  the remainder of the
molecule in determining the IR extinction
coefficient of the compound  at the
wavelength of interest.
  For example, three of the four aromatic
compounds tested have  reasonable re-
sponse factors «5) at 6.35 urn as shown
below. This wavelength is within a broad
aromatic ring stretch area.
     Compound
Response Factor
     Range
 Diisopropyl Benzene   2.42  - 3.75
 Dimethyl
   Styrene,2,4-

 Methyl Styrene
 0.185-0.394

 0.229-0.718
                   to incomplete resolution. Also, some shift
                   of the C-H stretch wavelength probably
                   occurs due to nearby oxygen or halogens.
                   A list of aliphatic compounds and corres-
                   ponding response factor ranges at this
                   wavelength is shown in Table 2. If four
                   alkylated  aromatic compounds are in-
                   cluded  in the  list  of compounds  with
                   response factors less than 5 at 3.3 - 3.4
                   um, the percentage of compounds tested
                   with suitable response factors increases
                   to 62%.
                      Ten chlorinated hydrocarbons tested in
                    this  program  yielded measurable re-
                    sponse factors at 13.5 um; 70% yielded
                    response factors less than 5 at this
                    wavelength.
                      Since the ultimate goal of this instru-
                    ment evaluation is to assess the suit-
                    ability of IR as a  general VOC screening
                    technique, an assessment of the useful-
                    ness of a single wavelength for measure-
                    ment of  organic  compounds of  varied
                    molecular weight and functionality is in
                    order. A review of the data indicates that
                    32 test compounds yield response factors
                    of 5.0 - 0.2 at each analytical wavelength:
                                           Number of
                      Wavelength (//m)     Compounds
                                         Table 2.    Substituted Aliphatic Compounds
                                                    with Response Factors Less than
                                                    20 at 3.3 yum
3.3
3.4
3.6
5.7
6.35
8.8
9.5
13.5
12
4
3
1
3
4
15
7
  Within this group, the addition of the
large aliphatic group (isopropyl) on the
benzene ring appears to reduce  the
sensitivity (larger response factor) at the
aromatic C — C stretch wavelength as
compared to less alkylated aromatics.
  For aliphatic and substituted aliphatic
compounds, the C-H stretch wavelength
of 3.3 jum yields suitable response factors
(<5) for about 52% of those tested. The
classical aliphatic C-H stretch is observed
at 3.4/um, but some overlap of 3.3 and 3.4
/Km IR bands may occur in the Miran due
•fr U S GOVERNMENT PRINTING OFFICE^ 1983-
   The results indicate that only 3.3, 9.5,
 and  13.5 /urn analytical  wavelengths
 respond acceptably for a large number of
 compounds. However, in any case, fewer
 than 50% of the compounds are reliably
 detected. Thus, there is  apparently no
 useful  agreement in response factors
 between, for example, a large number of
 aromatic compounds and aliphatic com-
 pounds (e.g., 50%  of  those tested)  at
 analytical wavelengths specific to each
 compound class. This observation indi-
 cates that infrared spectrophotometry is
 not particularly suitable for general VOC
 screening.
   On the other hand, the fact that the
 response factors  do not  vary by large
 values  (i.e.,  greater than  5.0) for some
 classes of compounds (e.g., halogenated
 aliphatics at 13.5/urn and aliphatic and
 alkylated aromatics at 3.3 -  3.4 /um)
 corroborates the  suitability of infrared
 spectrophotometry for VOC screening  of
 compounds belonging to one functional
 group.
- 659-O17/O89B
                                                Compound
                         Response
                        Factor Flange
                                         Acetyl-1 -propanol,3-
                                         Chloro-acetaldehyde
                                         Dichloro-1 -propanol.2,3-
                                         Dichloro-2-propanol, 1,3-
                                         Diketene
                                         Dimethylsulfide
                                         Ethanol
                                         Ethyleneglycoldimethyl
                                           Ether
                                         Ethyleneglycolmonoethyl
                                           Ether Acetate
                                         Formaldehyde
                                         Formic Acid
                                         Glycidol
                                         Methanol
                                         Methylene Chloride
                                         Pentanethiol. 1 •
                                         Propylene Chlorohydrin
                                         Tetrachloroethane, 1.1.2,2-
                                         Trichloroethane. 1.1,1-
                       1.23  - 2.02
                       2.73  - 8.62
                          18.5
                           5.29
                       8.06  -14.1
                       0.488- 0.495
                       0.261 - 0.292

                       0.196- 0.296

                       0.280- 0.488
                       1.09  • 1.88
                       0.529- 0.722
                           0.382
                       0.294- 0.410
                       2.67  - 2.87
                       0.314- 0.633
                       0.334- 0.4Q3
                       8.59  - 9.90
                       1.69  - 3.76
Conclusions
  In summary, based on the results of
this evaluation, it appears that:
1.  Infrared (IR) spectrophotometry is not
    suitable for general VOC screening,
    except for analysis of VOC emissions
    of a single organic functional group
    character.
2.  IR screening of organic compounds
    of a single functional class (e.g., C-
    C1) may be suitable for as many as
    80% of compounds in the class.
3.  IR screening at a wavelength corres-
    ponding to  both aliphatic  and aro-
    matic stretches may be suitable for
    as many as 50% of organic  com-
    pounds.
4.  A portable photoionization device is
    not  available for VOC screening at
    concentrations of  100  - 10,000
    ppmv.
5.  Development  of a reliable dilution
    probe for use  on a photoionization
    device is readily achievable.
6.  With such a dilution probe, it appears
    that a photoionization device with an
    11.7 or 11.8 eV UV lamp may be, used
    for reliable analysis of VOC fugitive
    emissions.

-------
     K. T, Menzies and R. E. Fasano are with Arthur D. Little, Inc., Cambridge, MA
       02140.
     Merrill D. Jackson is the EPA Project Officer (see below).
     The complete report, entitled "Evaluation of Potential VOC Screening Instru-
       ments, " (Order No. PB 83-139 733; Cost: $11.50. subject to change) will be
       available only from:
             National Technical Information Service
             5285 Port Royal Road
             Springfield, VA 22161
             Telephone: 703-487-4650
     The EPA Project  Officer can be contacted at:
             Industrial Environmental Research Laboratory
             U.S. Environmental Protection Agency
             Research Triangle Park, NC27711
United States
Environmental Protection
Agency
Center for Environmental Research
Information
Cincinnati OH 45268
Postage and
Fees Paid
Environmental
Protection
Agency
EPA 335
Official Business
Penalty for Private Use $300

-------