I
I
           United States      Office of Air Quality       EPA-450/2-78-041
           Environmental Protection  Planning and Standards     OAQPS No. 1.2-115
           Agency         Research Triangle Park NC 27711  October 1978
           _
           Guideline  Series
           Measurement of
           Volatile Organic
|          Compounds
I
I
I
I
I
I

-------
I
•                                     EPA-450/2-78-041
|                                    OAQPS No. 1.2-115
I
I
         Measurement of Volatile
•           Organic  Compounds
I
I
I
I
I                   Emissions Measurement Branch
                Emission Standards and Engineering Division
I
I
I
•               U.S. ENVIRONMENTAL PROTECTION AGENCY
                   Office of Air, Noise, and Radiation
•                Office of Air Quality Planning and Standards
                Research Triangle Park, North Carolina 27711
_                       October 1978

-------
                                                                                               I



                                                                                               I



                                                                                               I



                                                                                               I
                                     OAQPS GUIDELINE SERIES

The guideline series of reports is being issued by the Office of Air Quality Planning and Standards (OAQPS)B>
provide information to state and local air pollution control agencies; for example, to provide guidance on the
acquisition and processing of air quality data and on the planning and analysis requisite for the mamtenance«f
air quality. Reports published in this series will be available- as supplies permit-from the Library Services Off»
(MD35), U.S. Environmental Protection Agency. Research Triangle Park, North Carolina 2771 1; or, for a nominal
fee, from the National Technical Information  Service, 5285 Port Royal Road, Springfield, Virginia 22161.




                                                                                               I



                                                                                               I


                                                                                               I



                                                                                               I
                                 Publication No. EPA-450/2-78-041
                                       (OAQPS No  1.2-115)
                                                n
I



I



I



I



I



I

-------
I

I                                           PREFACE
«            Emphasis  on  the control  of volatile  organic  compounds  through  the
         State Implementation Plans,  new source  performance  standards,  and
I       national  emission standards  for hazardous air  pollutants  has
         created a need for standardized test procedures.   In  setting national
|       performance standards for new sources and national  emission standards
_       for hazardous  air pollutants, the Environmental  Protection  Agency has
™       followed a policy of establishing a  reference  method  for  each  regulated
•       source category and pollutant.   Under the State  Implementation Plan
         process, however, test methods  and erocedures  are defined by the States.
|       Thus, the case-by-case approach used by the  Environmental Protection
         Agency for national standards could  conflict with State established
•       methods.  In addition, the case-by-case approach  does not pro-
•       vide sufficient guidance to  the States  in their  efforts to  develop
         regulations for a large number  of sources and  organic compounds.
•            The purpose  of this document, therefore,  is  to provide guidance
         to the States  on  the measurement of  volatile organic  compounds from a
•       diversity of sources and pollutants  that  is  consistent with the methodology
•       being applied  by  the Environmental  Protection  Agency  as it  develops
         regulations for specific sources and pollutants.
I

I

I

I

I

-------
                                                                           I
                                CONTENTS

                                                                  Page No. •

METHODS FOR DETERMINING VOLATILE ORGANIC COMPOUNDS                         "
  AS CARBON IN STATIONARY SOURCES	    1

     INTRODUCTION	    1    1

     RETIONALE FOR SELECTING ORGANIC CARBON	    1    •

     RECOMMENDED REFERENCE METHOD 	    6

     ALTERNATE METHODS 	    8    I

     SCREENING METHODS 	    9

     REGULATORY LANGUAGE 	   10    I

ATTACHMENT 1.   REFERENCE METHOD FOR DETERMINATION OF TOTAL                 •
  GASEOUS NONMETHANE ORGANIC EMISSIONS AS CARBON - AUTOMATED               I
  ANALYZER VERSION	   11    ™

     1.  Principle and Applicability	   11    •
     2.  Range and Sensitivity	   11    •
     3.  Interferences	   11
     4.  Apparatus	   11    •
     5.  Reagents	   12    |
     6.  Analyzer Performance Specifications	   14
     7.  Procedure	   15    •
     8.  Calculations	;	   17    •
     9.  References	   18    *

ATTACHMENT 2.   DETERMINATION OF TOTAL GASEOUS NONMETHANE ORGANIC           •
  EMISSIONS AS CARBON:  MANUAL SAMPLING AND ANALYSIS PROCEDURE	   19    •

     1.  Principle and Applicability	   19    •
     2.  Apparatus	   19    |
     3.  Reagents	   28
     4.  Procedure	   30    M
     5.  Calculations	   38    •
     6.  Bibliography	   40

ATTACHMENT 3.   ALTERNATE TEST METHOD FOR DIRECT MEASUREMENT OF             •
  TOTAL GASEOUS ORGANIC COMPOUNDS USING A FLAME IONIZATION ANALYZER. 42    I

     1.  Principle and Applicability	 42    •
     2.  Range and Sensitivity	 43    |
     3.  Interferences	 43
     4.  Apparatus	 45    _
     5.  Reagents	 46    •
                                                                           I

-------
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
                                   CONTENTS (CONTINUED)
                                                                            Page No.

         ATTACHMENT 3.

•            6.   System Performance Specifications	    48
              7.   Procedure	    49
•            8.   Calculations	    52
              9.  References	     54

         TECHNICAL REPORT DATA SHEET	     55

-------
I

I                   METHODS  FOR  DETERMINING VOLATILE ORGANIC COMPOUNDS
                               AS CARBON  IN STATIONARY SOURCES
•       Introduction
              Volatile  organic  compound  (VOC) emission control regulations are
•       being developed  by EPA and  by State and local agencies to meet the
•j       oxidant control  needs.  In  some  cases, the regulations are in
         terms of the volatile  organic content of solvents.   In other cases,
I       they cover  organic volume or mass concentrations, mass emission
         rates, and  control equipment efficiencies.  Regardless of the
I       approach taken in the  regulation, consideration must be given to the
_       expression  of  emission limits in terms of what can be measured, and
™       to the cost and  practicality of  the test methods.
I            One concept of  volatile organic emission measurement is
         the determination of organic carbon mass concentration.
|       The rationale  for selecting this concept and conceptual approach
         for writing regulations  in  terms of volatile organic carbon are dis-
•       cussed herein, and two specific  test methods are presented to implement
•       the recommended  approach.
         Rationale for  Selecting  Organic  Carbon
I            In considering  volatile organic compound test methods one must
         recognize that organic emissions normally consist of a
•       mixture of  compounds and that there is presently no detection
•       technique having an  inherent, quantitative response to the
         total  molecular  structure of the mixture.  Several detection
•       techniques  respond to  organic compounds; however, the response
         can vary widely  from compound to compound and may, there-
I
fore, not be proportional  to the total  organic  mass  or  volume  in
a mixture.

-------
I
I
                                      2
     Such is the case of the  flame  ionization detector (FID), the
most commonly used detector for  organic measurements.  The FID  re-         •
sponse can vary from compound to compound because it is a function          •
of the number of carbon  atoms, the  type of bonds, and the elements
present in the organic molecules.  Thus, if the volatile organic            I
emission limit is expressed in terms  that require the measurement
of the total molecular structure of the organic emissions, the vari-        •
able response of the flame ionization detector must be overcome by          •
one of the methods described  below.
     1.  Gas Chromatograph/Flame Ionization Detector.  This method          I
         involves the separation of the organic components into dis-
         crete compounds using gas  chromatography (GC).  The compounds      8
         are identified, and  the FID  is calibrated for each of the          •
         identified compounds.  The compounds are then measured
         individually, and the total  mass concentration is determined       •
         by adding the individual mass concentration values; methane
         can be identified and excluded from the results.  This             |
         method may be eractical  where only two or three compounds are      •
         emitted, such as in  maleic anhydride plants; but if it is
         applied to sources that emit numerous organics, the time           fl
         and expense would be formidable.  For example, over 20
         peaks were noted in  a preliminary study of emissions from the      |
         manufacture of  nitrobenzene.                                      .
     2.  Direct Flame Ionization Detector with Emission Stream Character!-  "
         zation.  This method involves direct measurement with an FID analyBr,

                                                                           I

-------
I
I
                                              3
                  with prior characterization  of  the  gas stream and knowledge
|                that the detector responds predictably to the organic
_                components in the stream.  If present, methane will, of
™                course, also be measured.
•                     In practice, this  method can be applied to the determina-
                  tion of the mass concentration  of the total molecular struc-
•                ture of the organic  emissions under the following limited con-
                  ditions:  (1) where  only one compound is known to exist, (2)
•                when the organic compounds consist  of only hydrogen and carbon,
•                (3) where the relative  percentage of the compounds is known or
                  can be determined, and  the FID  response to the compounds is
•                known; (4) where a consistent mixture of compounds exists be-
                  fore and after emission control and only the relative concen-
•                trations are to be assessed, or (5) where the FID can be cali-
•                brated against mass  standards of the emissions (solvent emis-
                  sions, for example).
•                     In the case of  volatile organic solvents, accurate measure-
                  ments by direct FID  analyzers without calibration with solvent
•                standards are seldom possible because these solvents are often a
•                mixture of multiple  unknown  compounds.  Even if the emissions
                  can be separated and identified using a GC, accurate determina-
•                tion of the average  FID response is often impractical.  In addi-
                  tion, the emissions  may be altered  as they pass through a control
I                device: for example,  they may be partially oxidized in an
•                incinerator or selectively retained in an adsorber.  In such cases
                  the measurement is more difficult and cannot be corrected with
I

-------
I
I
                                    4
         solvent standards; therefore even the determination  of per-
         cent control  efficiency can be a difficult problem.                 I
     Another applicable measurement technique involves  an  oxida-
tion-reduction analysis; however,  the results of this  technique            |
are in terms of organic carbon and not a mass concentration of the           «
total  molecular structure.   In this approach, the nonmethane  organic
compounds are separated from other carbon compounds and are then            fl
oxidized to COp.  The  resultant CCL is subsequently reduced to
methane, which is then measured with an FID.   The C0?  from the              |
combustion step can also be measured with a nondispersive                    _
infrared (NDIR) analyzer; however, the NDIR is not as  sensi-                 "
tive as the FID and is therefore limited to high concentration              •
levels.  One limitation to  the oxidation-reduction analysis
is that the equipment  required is somewhat complex and  is                    I
unlikely to be made available in a portable form.
     Consideration of the various measurement approaches                    •
indicates that organic emission regulations expressed  in                    •
terms  of the measurement of organic carbon could be applied
to a wide range of volatile compounds.  The measurement of                  I
organic carbon can be  used  to assess directly the efficiency  of con-
trol devices such as incinerators or adsorbers.  By performing volume-       •
trie flow rate measurements, one can then determine organic carbon  emis-     •
sion rates.  Organic carbon content can also be related to volatile
I
I
I

-------
I
I
                                              5

         content of solvents  or surface  coatings.

I            Costs, logistics, and  other  practicalities of source testing

M       may, under limited conditions,  make  other test methods more

*       desirable for routine  compliance  determinations.  Three distinct

I       categories of test methods  are  therefore recommended for use with

         volatile organic  compound regulations expressed in terms of organic

J       carbon:  a reference method,  alternate methods, and screening methods,

_       These categories  are described  as  follows:

™            1.  Reference Method.  This  method would be applicable to

•                all  regulated sources  and would be accurate in reference

                  to the emission standard.   OAQPS recommends that the

•                reference method be  based on the oxidation-reduction

                  method of analysis to  measure organic carbon.

"            2.  Alternate Methods.   These are methods not necessarily

•                demonstrated  to be equivalent to the reference method,

                  but demonstrated to  the  satisfaction of the control

•                agency to produce  results adequate for determining com-

                  pliance, in specific applications.  Methods involving

•                direct measurement with  flame ionization detectors would

•                be primary  candidates  for alternate methods.

             3.   Screening Methods.   These alternative methods may produce

•                biased or imprecise  results, but they have been demonstrated

                  to the satisfaction  of the  control agency to be

|                adequate for  determining  compliance, provided that
I

I
                  any bias  or  imprecision  is  taken  into account.  These

-------
                                                                           I

                                                                           I
         methods are normally characterized  by  portability of               .
         equipment, procedural  simplicity, and  low cost.  Methods           *
         based on thermal  conductivity  or low-cost portable FID            •
         analyzers would be candidates for approval  as  screening methods.
Recommended Reference Method                                               p
     A reference method that involves  indirect  measurement of  volatile      —
organic carbon by an oxidation-reduction is  recommended.  This  neces-       •
sitates that the emission  limits  be expressed in  terms  of organic           •
carbon.  If the emission limits and a  universal reference method are
both based on organic carbon, the volatile organic standard will be         I
expressed in clear, unambiguous terms.  No other  known  practical test
method could accomplish this objective for a wide range of volatile         •
organic compounds.                                                         •
     A draft "Reference Method for Determination  of Total Gaseous
Nonmethane Organic Emissions as Carbon" is included as  Attachment  1.        •
The method requires a system for  separating  total  nonmethane organics
from other carbon compounds, converting the  total  nonmethane organics       •
to methane, and analysis of the methane by a flame ionization  detector.     •
Other than requiring this  general equipment, the  method provides per-
formance specifications designed  to assure correct performance  of           •
the separation-detection system.   Depending  on  the organic carbon
concentration, the method  may allow the use  of  the NDIR to detect           I
the CO,, formed by the initial oxidation step.                               •
     The concept upon which the method is based has been  utilized
for many years in Los Angeles County,  where  it  has been demon-              I
strated to be valid and effective for  compliance  determinations.
                                                                           I

-------
I
I
                                            7
         There  is  at  least one  private laboratory that is able to perform
'       the  analysis on  a fee  basis.  The oxidation-reduction analysis
•       concept for  measurement of gaseous nonmethane organic compounds
         is also offered  commercially by at least one instrument vendor.
I       Although  the availability of instruments is admittedly very
         limited,  there is no serious technical impediment that would pre-
•       vent additional  vendors from designing and producing acceptable
•       instruments.
             Because of  the somewhat limited use potential for the Los
•       Angeles laboratory-oriented procedure and the present limited
         production of commerical instruments for field use, the OAQPS
I       recommendation of the  oxidation-reduction reference method
•       for  the definition of  organic emissions is recognized as
         leading the  technology.  Wide acceptance of the organic carbon
•       reference method will, however, procide the needed inducement
         for  additional vendors to enter the market and thereby increase
|       the  supply and variety of organic carbon analysis instruments.
•           A draft of  the Los Angeles procedure is included as Attach-
         ment 2.   Although the  detail is considerable, some agencies may
I       wish to assemble the laboratory apparatus.  The Emission Measure-
         ment Branch  of OAQPS is working to refine the operating details
|       of the procedure; as information becomes available, it will be
         incorporated into the method.
I
I
I

-------
Alternate Methods
                                                                           I

                                                                           I

     An "Alternate method" may have only limited  applicability              I
or other deficiencies that prevent its  designation  as  a  reference           •
method.  Such methods may offer practical  advantages and produce
results adequate for determining compliance  in  certain applications.        I
Where such methods are applicable, the  results  may  be  accepted  in
lieu of reference method results.                                          I
     As a specific example, methods based  on direct measurement with        M
flame ionization detectors are often practical  for  hydrocarbon
compounds, and, in addition, the equipment is widely available.            Ij
An FID analyzer will be less costly and may  be  less complicated
for field application that an oxidation-reduction analyzer;  therefore,      |
for those applications where such methods  can be  made  to produce            .
accurate results, approval of them as alternate methods  is                  "
desirable.                                                                 •
     The Office of Air Quality Planning and  Standards, EPA,has  drafted
an "Alternate Test Method for Direct Measurement  of Total Gaseous  Organic   |
Compounds Using a Flame Ionization Detector," which is included as
Attachment 3.  This method outlines the known characteristics and  limi-     •
tations of FID techniques and provides  procedures needed to  assure its      •
proper operation.  The method does not  and is not intended  to
indicate specific applications where the method can (or  cannot) be          •
used or correction factors to be applied to  the results. Such  deter-
minations must be made on a case-by-case basis  founded on knowledge of       •

                                                                            I

                                                                            I

-------
I
I
I
                                             9
          the contents of the stream under test and the limitations of the
          detector.
•        Screening Methods
               In  addition to the limitations associated with alternate
|        methods, screening methods may also lack precision.  In spite
_        of such  shortcomings, screening methods may play an
•        important role in any volatile organic control program.  As a practical
•        matter,  the cost of applying a reference or alternate method to all
          or even  a majority of the regulated effluent streams in a jurisdic-
•        tion  may be unreasonable; therefore, less expensive, simpler testing
          techniques will be needed.
•             To  date, OAQPS has made effective use of an explosimeter to detect
•        vapor leaks in gasoline marketing operations.  In addition, an inexpensive
          hydrocarbon monitor using a solid-state ionization detector was designed
I        by OAQPS and has been used successfully as an emissions breakthrough
          detector on the exit of a carbon adsorber.  More recently, OAQPS has
•        initiated a test program associated with the development of new source
•        standards, using portable analyzers to detect leaks occurring in unit
          operations in the petroleum industry.  Two analyzers will be employed
•        in this  program, one involving a combustion/thermal-conductivity-type
          detector and the other a low-cost FID.
|             Another example of a screening method would be the case of an FID
«        analyzer applied to an unknown gas stream.  In such case there is often
          enough information available to provide a rough estimate of the analyzer
I        accuracy, but a more exact determination would be prohibitive. In such an

I

-------
                                                                            I
                                                                            I
event the FID may be used  as  an  alternative method  for determining           H
compliance, provided that  sufficient  buffer is  included  to account
for the possible inaccuracy.                                                 I
Regulatory Language
     Examples of how general  regulations  may  be expressed in terms of        |
the reference method that  measures  organic carbon concentration  are          _
as follows:                                                                 •
     1.  To regulate concentration:                                         •
         "Emissions of organic carbon shall not exceed 	
         grams carbon per  cubic  meter."                                      |
     2.  To regulate mass  rate:                                              _
         "Emissions of organic carbon shall not exceed 	              *
         grams carbon per  hour"  or  "Emissions of organic car-                •
         bon shall not exceed 	grams carbon per kilogram
         of solvent used."                                                  I
     To protect the analytical instrument from  contamination from
particulates and condensation, a filter  and heated  sample line (tern-         •
perature defined) must also be included  in the  emission  regulations.         •
                                                                            I
                                                                            I
                                                                            I
                                                                            I
                                                                            I

-------
•                                           n                            8/29/77

•           ATTACHMENT 1.  REFERENCE METHOD FOR DETERMINATION OF TOTAL GASEOUS
             NONMETHANE ORGANIC EMISSIONS AS CARBON—AUTOMATED ANALYZER VERSION
              1.   Pri nci pie and Appl i cab i 1 i ty
•                 1.1   Principle.   Conditioned stack gas  is  transported  to and
              analyzed  by a semiportable gas chromatograph (GC)  equipped  with  a flame
I            ionization detector (FID).  The total  gaseous nonmethane  organic (TGNMO)
I
fraction is separated by means of various GC columns from the other
constituents, oxidized to C02 and then reduced to methane (CH.)  before
I            it is introduced to the FID.   In  this  manner,  the  variable  response  of
              the FID associated with different types  of organics  is  eliminated, and  a
|            count of TGNMO carbon  atoms  is obtained.
M                 1.2  Applicability.   The method  is  applicable to the semicontinuous
              measurement of total gaseous  nonmethane  organics in  source  emissions.
V            2.  Range and Sensitivity
                   2.1  Range.  Signal  attenuators  shall  be  available so  that  a
I            minimum signal  response of 10 percent  of full  scale  can be  produced
•            when analyzing calibration gas or sample.
                   2.2  Sensitivity.   The  detector  sensitivity shall  be equal  to or
•            better than 2.0 percent of the full scale  setting, with a minimum full
              scale setting of 10 ppm (methane  or carbon  equivalent).
|            3.  Interferences
—                 None.
™            4.  Apparatus
•                 4.1   TGNMO analyzers are available  commercially or can  be constructed
              from available components by  a qualified instrument  laboratory.  The
I
I

-------
                                                                        I
                                 12
primary components of the analyzer are an FID preceded by a
GC column to achieve the necessary separation of TGNMO from other       I
carbon compounds.  Oxidation and reduction catalysts then convert
the TGNMO to CH. prior to detection.   The analyzer shall  be             |
accompanied by an instruction manual  (supplied by the manufacturer      _
if the analyzer was commercially produced) describing proper operation
and maintenance procedures.   In addition to the specific  procedures      •
required by this method, the analyzer shall be demonstrated prior to
initial use to be capable of proper separation, oxidation, and          |
reduction.  As a minimum this demonstration shall include measurement   _
of a known TGNMO concentration present in a mixture that  also contains   •
similar amounts of CH., CO^s and CO.   Certification of such demon-      •
stration by the manufacturer is acceptable.
     4.2  Sample Conditioning or Interface System (see Figure 1).       •
Probe with filter, 6.4 mm O.D.  Teflon  sample line, Teflon-coated
                                                                        I
                                                                        I
diaphragm pump, and Teflon flow control  valves.   A heating system
capable of maintaining all components at 120°C or greater shall  be
included.  The pump shall  be sized so that the sample residence  time
from the probe to the instrument will not exceed 15 seconds.             •
     4.3  Potentiometric Recorder (optional).   Strip chart recorder
with a voltage output compatible with the analyzer.                     •
5.  Reagents                                                            •
     5.1  Combustion Gas.   Air containing less than 2 ppm organics
(methane or carbon equivalent).                                         I
 Mention of trade names on specific products does not constitute        I
 endorsement by the Environmental  Protection Agency.                     *
                                                                        I

-------
1
HEATED 13
1 PROBE (~ 100 °C) HEATED
. TEFLON
FILTER x / / SAMPLE LINE
[GI^S WOOL) JX / (-100 °C)
• V-.J±L^L__ /
• cc4;v -----_-_' -~hr= — -/
1 xl
• STACK ' ///
WALL ^X'
IHEAT
TEFLON-C
DIAPHRAGM
(LEAKLE
1
1
HEATED TGNMO
•ANALYZER v
(~100°C) \
1 \
1
1
1
•
	 ' 'N\
l^oYi
ED ^^^ A^-V /
DATED ^^ dL/~\\, /
PUMP ' s~^
*t-. •?
^_I3
r 	 	 Jr^ —
CHROMATOGRAPHIC
SEPARATION
TGNMO
Optional j-
(CO,C02,CH^ OXIDATION
i CATALYST
r— -L_ , T
REDUCTION ] 1
CATALYST I C09
^nr r
V REDUCTION
CH CATALYST
4
\ 	 ^"4
FLAME 10N1ZATION
DL'TECTOR
t Optional
RESULTS - ^0,C09,CH/)TGf?MO
HEATED
FLOW CONT
/ VALVE
/ (~-100°C
' EX
/ B
'^
~=^=-^y=3
-* 	 SPAN GASES
^~ ZEKO GAS
"* CARRIER GAS
-«- 	 HYDROGEN
-<- 	 AIR
FIGURE 1. On-site Application of TGNMO Analyzer
1
EXCESS SAMP
 BLEED VALV

-------
                                                                       I
                                                                       I
     5.2  Fuel.  Hydrogen or a mixture of hydrogen and inert gas
containing less than 1  ppm organics (methane or carbon equivalent).
     5.3  Carrier Gas.   Helium, nitrogen, air or hydrogen  containing
less than 1 ppm organics (methane or carbon equivalent).
     5.4  Zero Gas.   Air containing less than 1  ppm organics (methane
or carbon equivalent).
     5.5  Calibration Gases (2).   Gas mixture standards with known      •
propane (C,HQ) concentrations corresponding to ranges  of 5-10 ppm and
5-10 percent (methane or carbon equivalent) are prepared and certified
by a gas manufacturer.   The mixture shall consist of CgHg,  CO,  C02,
and CH, in nitrogen.  The gas manufacturer must recommend  a maximum     •
shelf life for each  cylinder so that the C-^Hg concentration does  not    •
change more than +^5 percent from its certified value.  The date  of
gas cylinder preparation, certified C3Hg, CO, C02, and CH^ concentratiop
and recommended maximum shelf life must be affixed to  the  cylinder
before shipment from the gas manufacturer to the buyer. These gas      •
mixture standards are to be used to prepare a chromatograph calibration*
curve as described in Section 7.2.
     5.6 Span Gas.  The calibration gas corresponding  to 5 to 10  percenB
(methane or carbon equivalent) is used to span the analyzer.
6.  Analyzer Performance Specifications                                8
     6.1  Linearity:  +_ 5 percent of the expected value for full  scale  •
settings up to the maximum percent absolute (methane or carbon
equivalent) calibration point.  The analyzer shall be  demonstrated prioB

                                                                       I

                                                                       I

-------
I
                                             15

            to initial use to meet this specification through a 5-point (minimum)
I          calibration.  There shall be at least one calibration point in each
            of the following ranges:  5-10, 50-100, 500-1,000, 5,000-10,000, and
•          50,000-100,000 ppm (methane or carbon equivalent).  Certification
•          of such demonstration by the manufacturer is acceptable.  An additional
            linearity performance check (see Section 7.2.1) must be made before
•          each use.
                 6.2  Zero Drift.  One percent full scale per test period.
I               6.3  Span Drift.  One percent full scale per test period.
•          7.  Procedure
                 7.1  Sampling
•               7.1.1  Assemble the system as shown in Figure 1.  Locate the
            analyzer in a suitable environment.  Take particular care that sample
8          will be introduced to the system under the same conditions of pressure
•          and flow rates as are used in calibration.  For specific operating
            instructions for the TGNMO analyzer, refer to the operation manual.
I               7.1.2  Adjust the sampling system and analyzer heating system to
            provide a minimum temperature of 120°C and allow the system to warm up.
8               7.1.3  Perform a leak check as follows before sampling:  Recheck
«          to confirm that all fittings are tight.  With the sample probe plugged,
            open the flow control valve and the excess sample bleed valve.  Use
I          leak detection fluid or immerse the tubing leading from the bleed valve
            in a jar of water to check that sample flow has ceased.  At the con-
I          elusion of the sampling tests, recheck for leaks.
I
I

-------
I
I
                               16
     7.1.4  Begin Actual  Sampling.   Set the  signal  attenuation
to yield a minimum response of 10 percent  of full  scale,  unless  the    •
stack concentration is less than 1  ppm. Adjust  the flow  and  bleed
valves to minimize sample line residence time.   Perform the analysis
a minimum of four times.   Report the average of  the final  four
readings.  The analyzer cycle time is normally 10  to 15 minutes.       •
     7.1.5  At the conclusion of the sampling tests, but  at least once •
every day, introduce zero and span gas to  the analyzer to determine
zero and span drifts.   If the analyzer has drifted beyond the allowable*
performance specification, the tests shall be considered  invalid.
     7.2  Calibration                                                  I
     7.2.1  Calibration Curve.  Maintain a record  of performance of    •
each item.  Determine  the linearity of the analyzer for TGNMO as
follows:  With the signal attenuation at the most  sensitive setting,   I
introduce zero gas and adjust the respectivezeroing controls  to
indicate a reading of less than 1 percent  of full  scale.   With the     |
signal attenuation at  the least sensitive  setting, introduce  the span  •
gas and adjust the span control to indicate  the  proper value  on  the
analyzer readout.  Repeat these two steps  until  adjustments are  no     •
longer necessary.  Calculate a predicted response  for the 5-10 ppm
calibration gas.  Introduce that calibration gas and note the value    |
obtained.  If this value is not within +5  percent  of its  predicted     _
value, then the analyzer may need repairs, or one  or both of  the      ™
calibration gases may need replacement.  In  any  event, this  linearity  •
performance specification shall be met before the  analyzer is placed
in actual use.
 I

-------
I
I
                   7.3  Catalyst Performance Check.   These checks  should  be
•            performed on a frequency established by the amount of use of the
              analyzer, and the nature of the organic emissions  to which  it  is
I            exposed.  To confirm that the oxidation catalyst is  functioning  in
•            the correct manner, the operator must turn off or  bypass  the reduction
              catalyst while operating the analyzer in an otherwise normal fashion.
•            If oxidation is adequate, the only gas that will then reach the
              detector will be C09, to which the FID has no response.   If responses
                                 I2
              are noted, then the oxidation catalyst must be replaced.  To confirm
I
              the operation of the reduction catalyst,  reverse the above procedure.

              If CC"  in the calibration gases is  not reduced to CH^ as  it should be,

I            then the reduction catalyst must be replaced.

              8.  Calculations

|                 8.1  Determine concentrations  of TGNMO (propane equivalent)

_            directly from the calibration curves.   Multiply this number by 3  to

™            obtain ppm TGNMO (methane or carbon equivalent).

•                 8.2  Conversion to mass concentration  values for TGNMO as

              carbon is made as follows:
                             13
                   mg TGNMO/m  as carbon = ppm TGNMO (methane or carbon equivalent)

                        x 0.499

•            where:

                   II ppm TGNMO (methane or carbon equivalent) = ~-  x   41'57 9"mo1e
                                                                10b         ill
I


I
                                         0.499 mg/m3 as carbon.

-------
                                18
                                                                        I
where:
     Molecular weight of carbon =12                                    •
     Standard conditions:  20°C, 1  atm.
9.  References                                                          •
     9.1  Albert E. Salo, Samuel Whitz,  and  Robert D.  MacPhee.           •
"Determination of Solvent Vapor Concentrations  by Total  Combustion
Analysis:  A Comparison of Infrared With Flame  lonization  Detectors."    I
Presented at the 68th Annual  Meeting of  the  Air Pollution  Control
Association, Boston, Ma.  Paper No. 75-33.2. June 15-20,  1975.          |
     9.2  Instruction Manual, Byron Model  401 Total  Emission  Analyzer,   «
Byron Instruments, Inc., 520  1/2 S. Harrington  Street, Raleigh,  N.C. 2760™.
                                                                        I

                                                                        I

                                                                        I

                                                                        I

                                                                        I

                                                                        I

                                                                        I

                                                                        I

                                                                        I

                                                                        I

-------
I
                                          19
I                 ATTACHMENT 2.   DETERMINATION OF  TOTAL  GASEOUS  NONMETHANE
™                    ORGANIC EMISSIONS  AS  CARBON:  MANUAL SAMPLING AND
                                      ANALYSIS PROCEDURE
             1.   Principle and Applicability
•                1.1   Principle.   An  emission  sample  is anisokinetically
             drawn from the stack  through a heated  filter and a chilled
•           condensate trap by means  of an evacuated  gas collection tank.
             Total gaseous non-methane organics (TGNMO) are determined by
•           combining the analytical  results obtained from independent
•           analyses  of the condensate  trap and evacuated tank fractions.
             After sampling is completed, the organic  contents of the
I           condensate trap are oxidized to carbon dioxide which is
             quantitatively collected  in an evacuated  vessel; a portion
I           of the carbon dioxide is  reduced to methane and measured by
•           a  flame ionization detector (FID).  A  portion of the sample
             collected in the gas  sampling  tank is  injected into  a gas
•           chromatographic (GC)  column to achieve separation of the
             nonmethane organics from  carbon monoxide, carbon dioxide
I           and methane; the nonmethane organics are  oxidized to carbon
M           dioxide,  reduced to methane, and measured by a FID.
                  1.2   Applicability.  This method  is  applicable  to the
I           measurement of total  gaseous nonmethane organics in  source
             emissions.
|           2.   Apparatus
_                2.1   General.  TGNMO sampling equipment can be  constructed
*           by a laboratory from  commercially  available components and
I
I

-------
                               20
I
I
components fabricated in a  machine shop.  The  primary com-
ponents of the sampling system are a  heated  filter, condensate          I
trap, flow control  system,  and gas sampling  tank.   (Figure 1).
The primary components of the analytical  system  are an                  |
oxidation system for recovery of the  sample  from the condensate         _
trap and a TGNMO analyzer.   The TGNMO analyzer is  a FID                 •
preceded by an oxidation catalyst, a  reduction catalyst, and            •
a GC column with backflush  capability (Figure  2).   The system
for the removal and conditioning of the  organics captured in            •
the condensate trap consists of a heat source, oxidation
catalyst, Non-Dispersive Infrared (NDIR)  analyzer  and an inter-         •
mediate gas collection tank (Figure 3).                                 •
     2.2  Sampling.
     2.2.1  Probe.   1/8" stainless steel  tubing  heated to               I
approximately 120°C.
     2.2.2  Filter Holder.   Stainless steel  with a stainless            I
steel or glass frit filter  support and a  Teflon  gasket.  The            •
holder design shall provide a positive seal  against leakage
from the outside or around  the filter.  The  holder shall be             I
attached at the outlet of the probe.
     2.2.3  Filter Heating  System. Any  heating  system capable          I
of maintaining a temperature around the  filter holder during            •
sampling of 120 + 14°C (248 + 25° F), or such other temperature
as specified by an applicable subpart of the standards or                I
approved by the Administrator for a particular application.
A temperature gauge capable of measuring temperature  to  within           •

                                                                        I

-------
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
                                  21
                                                                              <
                                                                              ts
                                                                              <•
                                                                              V),
                                                                               Ol
  _
u _i

-------
                                     22
                SAMPLE
o
3
cj

<
a.
<
K

K
111
C/l
Z
LU
a
z
o
o
cc
o
u_

K
<
«V
>-
CO
z
S
_j
o
o
                                 <3
                                          CARRIER GAS
                              SEPARATION
                               COLUMN
                  CONTROL VALVE ( ^
                               CO
                               I
                               CH4
                               C02
                  CONTROL VALVE
                            OXIDIZING
                            CATALYST
                               i
                              WATER
                              TRAP
 BACKFLUSH
• CARRIER
    GAS
                  TGNMO
                BACKFLUSH
                                   CONTROL VALVE
                            REDUCTION
                            CATALYST
               FID
                                                     1     f
                                         COMBUSTION AIR
                 FUEL
Figure  2.   TOTAL GASEOUSNONMETHANE ORGANIC (TGNMO) ANALYZER SCHEMATIC
                                         I
                                         I
                                         I
                                         I
                                         I
                                         I
I
I
I
I
I
I
I
I
I
I
I

-------
I
I
I
I
I
I
I
I
I
I
I
I
I
 I
 I
I
 I
 I
23
                DC
                O
                        <0
                        O t-
                                                                     <
                                                                     ec
                                                                     <
                                                                     C9
                                                                     Z
                                                                     O
                                                                     Z
                                                                     O
                                                                     o
                                                                     O

                                                                     <

                                                                     ec
                                                                     ui
                                                                     >
                                                                     o
                                                                     o
                                                                     Ul
                                                                     ec
                                                                     a
                                                                     Z
                                                                     o
                                                                     o
                                                                     O)
                                                                CO
                                                             o
                                                             u.
                                                             *
                                                                u
                                                                u.
                                                                o

-------
                            24

3° C (5.4° F) shall  be installed so that the  temperature
I
I
I
around the filter holder can be regulated  and  monitored
during sampling.                                                        •
     2.2.4  Condensate Trap.  The condensate trap shall  be
constructed of 316 stainless steel;  construction  details of a            •
suitable trap are shown in Figure 4.
     2.2.5  Flow Control System.                                        •
     2.2.5.1  Needle Valve.  To regulate sample gas  flow rate.           •
     2.2.5.2  Rate Meter.  Rotameter,  or equivalent  capable
of measuring flow rate to within +_10  percent  of  the                    I
selected flow rate of about 80 cc/min. Other flow control systems
capable of maintaining a constant sample rate  of  80  cc/min  +_            •
10 percent may be used subject to the  approval  of the Administrator.     •
     2.2.6  Gas Collection Tank.  Stainless steel or aluminum  tank
with a minimum volume of 6 liters.  The tank is fitted with a            I
vacuum gauge, a leak!ess valve, and a  t-connector for conducting
leak checks.                                                            I
     2.3  Analysis.  For analysis, the following  equipment  is  needed.    •
     2.3.1  Condensate Recovery and Conditioning  Apparatus
(Figure 3).                                                             I
     2.3.1.1  Heat Source.  A heat source  sufficient to  heat
the condansate trap to a "cherry red"  color.  An  electric                j|
muffle-type furnace or bunsen burner may be used.                       _
     2.3.1.2  Oxidizing Catalyst.  A platinum  and quartz                ™
catalyst constructed from a 44-inch length of  1/4" tubing of            I

                                                                        I

-------
 I
 I
 I
 I
 I
 I
 I
 I
 I
 I
                      25
            CONNECTOR
EXIT TUBE. 6mm (% in) 0.0. —
 X 0.71mm (0.028 in) WALL
6mm EXIT TUBE CONTAINS 3mm (1/8 in) SS WOOL PLUGS
OIMITHER SIDE OF A 6mm ('/. in) QUARTZ WOOL FILTER
  I

  I

  I

  I

  I

  I


  I

  I
      N0.40HOLE
        WELDED JOINTS
                                 INLET TUBE, 6mm (% in) O.D.
                                  X 0.71 mm (0.028 in) WALL
                                                    CONNECTOR
                                         	CRIMPED AND WELDED GAS-TIGHT SEAL
                                            BARREL 19mm (% in) O.D. X 140mm (5-'/i in) LONG
                                                    1.5mm (1/16 in) WALL
                                           BARREL PACKING. SS WOOL PACKED TIGHTLY
                                                 AT BOTTOM, LOOSELY AT TOP
                                             HEAT SINK (NUT, PRESS-FIT TO BARREL)
                                       WELDED PLUG
              MATERIAL: TYPE 316 STAINLESS STEEL
         Figure 4   CONDENSATETRAPZ

-------
                                                                       I
                             26
                                                                       I
70 percent Ni-30 percent Cr  alloy packed as follows:
     First 4 inches—empty.                                             I
     Next 4 inches--8-10 mesh alumina  coated with 0.5 percent           •
finely divided platinum.
     Next 28 inches—8 mesh  quartz chips.                               I
     Last 8 inches—8-10 mesh alumina  0.5  percent platinum coated.
     Other catalyst systems  capable  of meeting  the catalyst             |
efficiency criteria of this  method (Section 4.4.2) may be used          _
subject to the approval  of the  Administrator.                           ™
     2.3.1.3  Water Trap. Any  leak  proof  moisture trap capable         •
of removing moisture from the gas stream may be used.  A
condensate trap designed according to  the  specifications of             |
Figure 4 without packing in  the exit tube  will  suffice.                 _
     2.3.1.4  NDIR Detector. Detector capable  of indicating the        •
COp level in the zero  to five percent  range; required to monitor        •
the combustion progress of the  organic matter in the condensate
trap.                                                                  I
     2.3.1.5  Pressure Regulator.  Stainless steel needle
valve required to maintain the  NDIR detector at a constant              I
pressure.                                                              •
     2.3.1.6  Intermediate Collection  Tank.  Stainless steel
or aluminum collection vessel.   Tanks  with nominal volumes of           I
2 and 6 liters are recommended.  The end  of the tank  is  fitted
with a t-connector, vacuum gauge, and  leakless  valve.                   I
     2.3.1.7  Calibration Injection Port.   Injection  port valve         •
and sample loop for injection of calibration standards required
                                                                       I
                                                                       I

-------
I
I
                                             27
               to check the combustion efficiency of the condensate recovery
 g             system.
                    2.3.2  Total  Gaseous  Nonmethane Organic  (TGNMO) Analyzer.
 •             Semicontinuous GC/FID analyzer capable of:   (1)  separating  CO,
 •             COp, CH., and gaseous nonmethane organics,  (2) oxidizing  the
               nonmethane organic fraction to CO,,,  reducing  the C02 to methane,
 •             and quantifying the methane.   The analyzer shall  be  demonstrated
               prior to initial  use to be capable of proper  separation,  oxidation,
 •             reduction, and measurement.  As a minimum,  this  demonstration
 •             shall include measurement  of a known TGNMO concentration  present
               in a mixture that also contains CH. , CO,  and  COp.   (see paragraph
 I             4.4.1)  In addition, the analyzer shall meet  the following  per-
               formance specifications:
 I                  2.3.2.1  Linearity.  j^5 percent of  the  expected value for
 •             each full scale setting up to the maximum percent absolute
               (methane or carbon equivalent) calibration  point.  The analyzer
 I             shall be demonstrated prior to initial  use  to meet this specifica-
               tion through a 5-point (minimum) calibration.  There shall  be at
 |             least one calibration point in each  of the  following ranges:
 •             5-10, 50-100, 500-1,000, 5,000-10,000,  and  40,000-100,000 ppm
               (methane or carbon equivalent).   Certification of such demonstra-
 I             tion by the manufacturer is acceptable.   An additional linearity
               performance check (see Section 4.4.1.1) must  be  made before each
|             use.
 «                  2.3.2.2  Zero Drift.   One percent  full scale per analysis of
               an emission test  series.
I
I

-------
                               28
     2.3.2.3  Span Drift.   One percent full  scale  per analysis of
     1
      Mention of trade names or specific products does not
I
I
an emission test series.                                                  I
     2.3.2.4  The following components  have  been  found  to  be
acceptable for use in the TGNMO System:                                   I
     2.3.2.4.1  Oxidation Catalyst.   Type  316  stainless steel             •
0.25 inch OD tubing x 14  inches long packed  with  Hopcalite 25  -
30 mesh; operated at 850° C.                                             •
     2.3.2.4.2  Reduction Catalyst.   Type  316  stainless steel
0.25 inch OD tubing x 7 inches long  packed with 10 percent nickel         |
on chromasorb W, 60-80 mesh; operated at 400°  C.   Method of pre-          .
paration:  100 grams chromasorb W,  10 grams  nickelous nitrate,
75 ml water, evaporated to dryness  then heated in air for  4              •
hours to convert to nickel oxide.   After packing  the tubing,
reduce overnight at 450°  C and 30 ml/min H2  to nickel metal.              |
     2.3.2.5.3  Separation Column.   Type 316 Stainless  steel              —
0.125 inch OD tubing x 18 feet long packed with Porapak Q  60/80          •
mesh; operated isothermally at 80°  F.                                    •
     2.3.3  Mercury Manometer.  U-tube  mercury manometer
capable of measuring pressure to within 1.0  mm Hg in the 0 -              g
900 mm range.                                                            _
     2.3.4  Barometer.  Mercury, aneroid,  or other barometer              •
capable of measuring atmospheric pressure  to within 2.5 mm               •
(0.1 inch Hg).
3.   Reagents                                                             J
     3.1  Sampling.
                                                                         I
 constitute endorsement by the EPA.                                        |
                                                                          I

-------
                                         29
I              3.1.1   Filter.   Glass  fiber  filter without organic binder,
•         exhibiting  at least  99.95  percent efficiency  (<_ 0.05 percent
           penetration) on 0.3  micron  dioctyl  phthalate  smoke particles.
I         The filter  efficiency test  shall  be conducted in accordance with
           ASTM standard method D 2986-71.   Test  data  from suppliers
I         quality control program are sufficient for  this purpose.
_              3.1.2   Crushed  Dry Ice.
"              3.2 Analysis.
•              3.2.1   (TGNMO)  Analyzer.
                3.2.1.1  Carrier Gas.   5  percent  Op  in N« containing less
I         than 1  ppm  organics.
                3.2.1.2  Fuel Gas. 40 percent hydrogen  in nitrogen con-
•         taining less than 1  ppm organics.
•              3.2.2   Condensate Recovery and Conditioning Apparatus.
                3.2.2.1  Carrier Gas.   5  percent  02  in N2 containing less
•         than 1  ppm  organics.
                3.2.2.2  Oxygen.  Oxygen  containing  less than 1 ppm organics,
•              3.3 Calibration.
•              3.3.1   (TGNMO)  Analyzer.
                3.3.1.1  Calibration  Gases (3).   Gas mixture standards with
•         known propane (C,H_) concentrations corresponding to ranges of
           5-10 ppm, 50-10 percent and 20-25 percent methane or carbon
•         equivalent  are prepared and certified  by  a  gas manufacturer.
•         The Mixture shall  consist  of C_Hfl,  CO, C0«, and CH. in nitrogen.
           The gas manufacturer must  recommend a  maximum shelf life for
•         each cylinder so that the  C-HR concentration  does not change
I
I

-------
                               30
more than +_ 5 percent from its  certified  value.   The  date of                  I
gas cylinder preparation,  certified C,Hg, CO,  CO-,  and  CH.  concen-            M
trations and recommended maximum shelf life must  be affixed to
the cylinder before shipment from the gas manufacturer  to the                 I
buyer.  These gas mixture  standards are to be  used  to prepare a
chromatograph calibration  curve as described in Section 4.4.1.1.              |
     3.3.1.2  Span Gas. The calibration  gas (Section 3.3.1.1)                _
corresponding to 20 to 25  percent is used to span the analyzer.               ™
     3.3.1.3  Oxidation Catalyst Check.  The calibration gas                  •
(Section 3.3.1.1) corresponding to 20 to  25 percent is  used to
check the oxidation catalyst.                                                g
     3.3.1.4  Reduction Catalyst Check.  A gas standard with a
known concentration of 5 percent (nominal) C0« in nitrogen  is                 *
used to check the reduction catalyst.                                        •
     3.3.2  Condensate Recovery and Conditioning  Apparatus. Gas
mixture standards (2) with known propane (C,Hg) concentrations                •
in nitrogen corresponding  to ranges of 5-10 ppm and 5-10 percent
(methane or carbon equivalent)  are prepared and certified  by a                •
gas manufacturer.  These gas mixture standards are  to be used  to              •
check the operation of the condensate trap oxidation  system as
described in Section 4.4.2.                                                  •
4.  Procedure
     4.1  Sampling                                                           I
     4.1.1  Pretest Preparation.  The sample tank shall be cali-              •
brated  according to the procedure described in paragraph 4.4.3.
Check filters visually against the light for irregularities,                  •

                                                                             I

                                                                             I

-------
I
I
              flaws, or  pinhole leaks.  Either in the laboratory or in the field
|            evacuate the  sample tank to a vacuum of 755 mm mercury (measured
              by  a  mercury  U-tube manometer).  Record the temperature, bara-
•            metric pressure, tank vacuum measured with the manometer, and
•            the vacuum indicated on the tank gauge.
                   4.1.2 Assemble the system as shown in Figure 1.  Immerse
•            the condensate trap in dry ice and start the filter and probe
              heaters.
•                 4.1.3 Leak check procedures.
•                 4.1.3.1  Gas Sampling Tank Leak Check.  Leak check the gas
              sampling tank immediately after the tank is evacuated.  Once
•            the tank is evacuated, allow the tank to sit for 30 minutes.  The
              tank  is acceptable if no change in tank vacuum (measured by the
I            mercury manometer) is noted.
•                 4.1.3.2  Pretest Leak Check.  A pretest leak check is
              recommended,  but not required.  If the tester opts to conduct
I            the pretest leak check, the following procedure is used.  After
              the sampling  train has been assembled (including cooling of
I            condensate trap and heating of filter) plug the probe tip.
H            Attach the vacuum line of the leak check apparatus (Figure 5)
              to  the T-connector of the evacuated tank; open the valve on this
V            connector  (not the sample flow control valve to the evacuated
              tank) and  evacuate the sample train to a vacuum of 625 mm Hg.
|            Shut  the valve on the pump side of the manometer and allow the
H            sampling train to sit for 10 minutes.  A leak rate in excess of
              0.5 mm Hg  for this 10 minute period is unacceptable.  When the
I

I

-------
                   32
          CONTROL
           VALVE
CONTROL
 VALVE
BYPASS
VALVE
VACUUM
  LINE
          MERCURY
         MANOMETER
                                             VACUUM
                                              PUMP
                        LEAK CHECK
           Figure 5.    APPARATUS
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
 I
I
 I
 I

-------

                                         33
              leak check is completed, slowly release the vacuum 1n the train
•            by unpluging the probe, close the T-connector valve,  and plug
              this connector to assure a leak free system.
|                 4.1.3.3  Post Test Leak Check.   A leak check is  mandatory
M            at the conclusion of each test run.   After sampling is completed,
              plug the end of the sampling probe and attach the vacuum line of
B            the leak-check apparatus (Figure 5)  to the evacuated  tank
              t-connector.  Assure that the flow valve to the evacuation pump
|            (valve between manometer and pump) is closed.  First  open the
_            t-connector valve to the manometer and then open the  flow control
•            valve to the evacuated tank.  Record the clock time and tank
•            vacuum.  After 10 minutes note the tank vacuum.  A leak rate in
              excess of 0.5 mm Hg for this ten minute period is unacceptable
I            and the sampling run shall be voided.  After completing the leak
              check, close the evacuated tank flow control valve and the
™            t-connector flow control valve.  Disconnect the leak  check
•            apparatus and plug the t-connector to assure a leakproof seal
              during shipping.  Unplug the probe tip.
j|         .        4.1.4  Sample Train Operation.   Place the probe  into the
              stack such that the probe tip is located at a pre-selected location.
•            For stacks under negative pressure,  assure that the sample port
H            is sufficiently sealed to prevent leakage of ambient  air around
              the probe.  Record the clock time, sample tank gauge  vacuum, and
•            barometric pressure.  Assure that the flow control  needle valve
              is closed.  Begin sampling by opening the evacuated tank flow
I            valve all  the way.  Open the flow control  needle valve until

I

I

-------
                           34
I
I
the rotameter indicates the desired  setting; maintain a constant
flow rate (+ 10 percent) throughout  the  duration of the sampling          •
period.  Record the gauge vacuum,  rotameter setting, and filter
temperature at 5 minute intervals.   Select a total sample time            |
greater than or equal  to the minimum sampling  time specified  in           m
the applicable subpart of the standard;  end the sampling when
this time period is reached or when  a constant flow rate can              fl
no longer be maintained.  When the sampling is completed, close
the evacuated tank valve and remove  the  probe  from the stack.             |
Record the final readings.  Conduct  the  post test leak check              _
according to the procedures of paragraph 4.1.3.3.                         ™
     If the sampling must be stopped before obtaining the minimum         •
sampling time specified in the applicable subpart because a
constant flow rate cannot be maintained, proceed as follows:              I
After removing the probe from the  stack, conduct the post
test leak check.  After the leak check is completed, remove the           •
evacuated tank from the sampling train (without disconnecting            •
other portions of the sampling train) and connect another
evacuated tank to the sampling train.  Proceed with the  sampling;         •
after the minimum total sampling time is exceeded, end the test.
      \.2  Sample Recovery.                                               •
     Disconnect the condensate trap  at the filter and at  the  flow         M
metering system.  Tightly seal the ends  of the condensate trap;
keep the trap packed in dry ice until analysis is conducted.   Seal        •
the connection at the evacuated tank to  assure a  leak  proof  seal
during shipping.  After the evacuated tank has cooled  to  ambient          I

                                                                         I

                                                                         I

-------
 I
                                         35
             conditions, attach the U-tube manometer to the t-connector,
 fl           open the valve, and record the tank vacuum, ambient  temperature,
             barometric pressure, and indicated gauge vacuum.   Close  the  flow
 |           valve and reseal the t-connector to assure a leak  proof  seal
 _           during shipping.  Assure that the test run number  is properly
 *           indentified on the condensate trap and evacuated tank(s).
 •                4.3  Analysis
                  4.3.1  TGNMO Analyzer.  Heat the catalysts to their operating
 |           temperatures and set the carrier gas and fuel  flow rates.  Conduct
 _           the calibration check required in paragraph 4.4.1.1  and  the
 •           catalyst performance checks required in paragraph  4.4.1.2  prior
 •           to analyzing the test samples.
                  4.3.2  Condensate Trap.  Return the condensate  trap to  the
 Jj           laboratory and hook it into the recovery and conditioning  system
 _           (Figure 3).  Set the oven for the oxidizing catalyst at  850° C
 •           and the trap heating furnace at 600° C.  Set the gas directing
 •           valve to permit flow of 5 percent 02/N2 through channel  A  to the
             condensate trap at a rate of 80 cc/min; at the same  time set the
 I           oxygen flow through channel B at 20 cc/min (1:4 ratio).  After
             two minutes, switch the gas directing valve to permit the  oxygen
 •           to flow via channel A directly through the condensate trap and
 •           the 5 percent CL/N2 carrier gas to flow through channel  B.  When
             the NDIR indicates that COp is no longer being emitted from  the
 •           combustion system, shut off the collection flask from the  system
             and cease combustion.  Record the collection flask pressure  after
 •           combustion is completed (P.) and then pressurize the flask to
 •           860 mm Hg (nominal) with nitrogen and record the final pressure

I

-------
                           36
record the tank vacuum (Pt).   Pressurize  the  tank with
nitrogen and record the final  tank pressure  (P.  },  temperature
                                                                        I
                                                                        I
(Pf).  Remove a syringe sample from the flask and inject  this
into the TGNMO analyzer.  Record the analyzer response (ppm C)           |
for triplicate samples.
     4.3.3  Gas Sampling Tank.  Using a U-tube mercury manometer,        •

                                                                        I
l<(W«V^J«*ll %A I I w I *• W 1 *•• ** I I W * I I IM t  VV( I I It h/ISaWMfWII^ II I  / J  1*^1111^^1 VI VMlb
                                               f                        •
and barometric pressure.  Remove a syringe sample from the  tank          •
and inject this into the TGNMO analyzer.  Record the analyzer
response (ppm C) for the non-methane organic fraction for triplicate    •
samples.                                                                •
     4.4  Calibration.   Maintain a record of performance  of
each item.                                                              I
     4.4.1  TGNMO Analyzer.
     4.4.1.1  Calibration Curve.  Determine the linearity of            •
the analyzer for TGNMO  as follows:  With the signal  attenuation          •
at the most sensitive setting, introduce zero gas and adjust
the respective zeroing  controls to indicate a reading of  less            I
than 1 percent of full  scale.   With the signal attenuation  at
the least sensitive setting, introduce the span gas  and adjust           •
the span control to indicate the proper value on the analyzer            m
readout.  Repeat these  two steps until adjustments are no longer
necessary.  Calculate a predicted  response for the 5-10 ppm             I
calibration gas.  Introduce that calibration gas and note the
value obtained.  If this value is  not within j^5 percent  of its          |
predicted value, then the analyzer may need repairs, or one or           •
both of the calibration gases may  need replacement.   In any
                                                                        I
                                                                        I

-------
I

I
              event, this linearity performance specification shall  be met
|            before the analyzer is placed in actual use.
_                 4.4.1.2  Catalyst Performance Check.  These checks should be
•            performed on a frequency established by the amount of use of the
•            analyzer and the nature of the organic emissions to which it is
              exposed.  To confirm that the oxidation catalyst is functioning in
•            the correct manner, the operator must turn off or bypass the reduction
              catalyst while operating the analyzer in an otherwise normal
•            fashion.  Inject the calibration gas (paragraph 3.3.1.3) into the
•            system.  If oxidation is adequate, the only gas that will then
              reach the detector will be C02» to which the FID has no response.
•            If a response is noted, the oxidation catalyst must be replaced.
              To confirm the proper operation of the reduction catalyst, inject
•            a sample of the C(L calibration gas (Section 3.3.1.4)  into the
•            system.  If the C(L is not reduced to CH. as it should be, then
              the reduction catalyst must be replaced or regenerated.
•                 4.4.2  Condensate Trap Oxidation Catalyst.  Inject syringe
              samples of the calibration gases listed in Section 3.3.2 into the
I            sample port of the condensate trap combustion system (Figure 3).
m            Proceed with a normal  analysis (i.e., collection of the CCL in
              the flask followed by analysis of triplicate aliquots  using
I            the TGNMO analyzer) and compare results to the actual  concentration.
              Repair the system if the results (average of triplicate aliquots)
|            deviate by greater than +_ 5 percent from the calibration gas value.
«                 4.4.3  Gas Sampling Tank.  The volume of the gas  sampling
              tanks used must be determined.  Prior to putting each  tank in
I

I

-------
                                                                            I
                              38
service, determine the  tank  volume  by weighing the tanks empty               |
and then filled with  water;  weigh to the nearest 0.5 gm and                  M
record the results.
     4.4.4  Intermediate  Collection Flask.  The volume of the inter-         M
mediate collection flasks used  to collect CO- during the analysis
of the condensate traps must be determined.  Prior to putting                |
each flask in service,  determine the volume by weighing the                  «
flasks empty and then filled with water; weigh to the nearest                "
0.5 gm and record the results.                                               •
     4.4.5  Condensate  Trap  Leak Check.  Prior to each use, check
each condensate trap  for  leaks  by pressurizing with N« to approximately      J
50 psig and immersing in  water.
     4.4.6  Rotameter.  The  rotameter  need not be calibrated but             •
should be cleaned and maintained according to the manufacturer's             •
instruction.
5.  Calculations                                                            I
     5.1  Sample Volume.   For each  test  run calculate the gas
volume sampled:                                                             •
                                     /pt    PII\                           *
                          Vs .  0.36 Vf ^ - T- j                           I
                                     \  *     ri '
     5.2  Noncondensible  TGNMO.  For  each  collection  tank,  determine         I
the concentration of TGNMO (ppm C):
                                                                            •
     5.3  Condensible TGNMO.  For each condensate trap determine the
concentration of TGNMO (ppm C):                                             •
                                                                            I
                                                                            I

-------
1

1
1
^B
1

1
1

1

1

1
|
•I
1

1



1





1
1
1

39
3 X V- X P. n
r - " T v v C
Cc ' Vs X P. x kf1 Lcpk
5.4 Total Gaseous Nonmethane Organics (TGNMO). To
determine the TGNMO concentration for each test run, use the
following equation:
ct = c + cc
5.5 Control Device Efficiency. To determine the TGNMO control
device efficiency for each test run, use the following equation:
r - r
r _ Lti LtO v inn
t - - V A iV/U
Cti
where:
C « Noncondensible TGNMO calculated concentration, ppm
carbon equivalent.
C = TGNMO analyzer measured concentration for gas
collection tank, ppm propane.
C = Condensible TGNMO (condensate trap) calculated concentra
tion, ppm carbon equivalent.
C = TGNMO analyzer measured concentration for intermediate
collection flask, ppm propane.

C. = Total gaseous nonmethane organic (TGNMO), ppm carbon
equivalent.
C. = TGNMO at control device outlet, ppm carbon equivalent.
C. . = TGNMO at control device inlet, ppm carbon equivalent.
E = Control device efficiency, percent.

Pf = Final pressure of intermediate collection flask
(nominal 860 mm Hg.), mm Kg, absolute.



-------
                                                                            I
                              40
                                                                            I
     P.    = Pressure of intermediate  collection  flask  at  completion          _
            of combustion,  mm Hg,  absolute.                                  ™
     Pt    = Gas sample tank pressure  prior  to  sampling, mm  Hg,               •
            absolute.
     P.    = Gas sample tank pressure  after  sampling, but  prior  to            Jj
            pressurizing, mm Hg,  absolute.
     P.    = Final  gas sample tank pressure  after pressurizing,               •
            mm Hg, absolute.                                                 •
     T.    = Gas sample tank temperature prior  to sampling,  °K.
     T.    = Gas sample tank temperature at  completion  of  sampling,           •

                                                                            I
     T.    = Gas sample tank temperature after  pressurizing, °K.              •
       f
                                         3
     V    = Gas collection  tank volume, M
                                                  -3
     Vf   = Intermediate collection tank volume, M
     V    = Gas volume sampled, dscm                                        •
     m    = Total  number of injections of non-condensible TGNMO
            during analysis (where j  = injection number,  1  .  .  . m)          I
     n    = Total  number of injections of condensible  TGNMO during           •
            analysis (where k = injection number, 1 .  .  . n)
     0.36 = 273°K/760 mm Hg                                                 I
     Standard Conditions =  Dry, 760 mm Hg,  273°K.
6.  Bibliography                                                             •
     6.1  Albert E. Salo, Samuel  Witz, and  Robert D.  MacPhee.   "Deter-       •
mination of Solvent Vapor Concentrations by Total Combustion Analysis:
A comparison of Infrared with Flame lonization Detectors."   Presented        I
at the 68th Annual Meeting  of the Air Pollution Control  Association,
Boston,  Ma.  Paper No. 75-33.2.  June 15-20, 1975.                           §
I
 I

-------
I
                                        41
I
_                6.2  Albert  E. Salo, William L. Oaks, Robert D. MacPhee.
™           "Measuring the Organic Carbon Content of Source Emissions for Air
•           Pollution Control."  Presented at the 67th Annual Meeting of the Air
I
I
I
I
I
I
I
I
I
I
I
I
I
I
             Pollution  Control Association, Denver, Colorado.  Paper No. 74-190
             June 9-13,  1974.

-------
                                                                 June  6,  19||
                              42
 ATTACHMENT 3.  ALTERNATE TEST METHOD FOR DIRECT MEASUREMENT OF  TOTAL       |
      GASEOUS ORGANIC COMPOUNDS USING A FLAME IONIZATION  ANALYZER
                                                                           I
                             INTRODUCTION
                Performance of this method should not be                    |
                attempted by persons  unfamiliar  with  the
                performance characteristics of the flame                    —
                ionization detector,  nor by those who are                  I
                unfamiliar with source sampling.                           ™
1.  Principle and Applicability                                            •
     1.1  Principle.   The sample is drawn from the source,  through
a heated sample line and glass fiber  filter to a flame ionization          I
analyzer (FIA).  Ions formed in the combustion of a specific hydro-
carbon compound in a FL - 0^ flame establish a current that is              •
proportional to the mass flow rate of that hydrocarbon to the              •
flame.  This current is collected at  two polarized electrodes,
and is read out on a potentiometric recorder and compared with a           I
calibration curve based on propane (C3Hg), or an organic  solvent,
as appropriate.  The results are reported as equivalents  of                I
methane (CH.) or carbon, or in terms  of an organic solvent.                •
     1.2  Applicability.  This method is applicable for the
determination of the true carbon mass concentration,  and/or an               •
indicated volume or mass concentration (expressed in terms of
carbon or of an assumed organic compound, e.g.,  methane equivalent)          •
of gaseous organic compounds present  in an emission stream.  It              m
can also be used to measure the mass  concentration of an  organic
solvent if stable mass standards of the solvent  can be generated.           I

                                                                            I
                                                                            I
                                                                            I

-------
I
I
_       The measurement will not exclude methane, so a supplemental
™       measurement of methane may be necessary.
•       2.  Range and Sensitivity
              2.1  Range.  Signal attenuators shall be available so that
•       a minimum signal response of 10 percent of full scale can be pro-
         duced when analyzing calibration gas or sample.
•            2.2  Sensitivity.  The detector sensitivity shall be equal
•       to or better than 2.0 percent of the full scale setting, with a
         minimum full scale setting of 10 ppm (methane or carbon equivalent)
I       3.  Interferences
              3.1  Nonorganic Gases.  There is no response to nitrogen,
•       carbon monoxide, carbon dioxide, or water vapor, however, the
•       analyzer response to organics will be affected by the composition
         of the background or carrier gas.  It is, therefore, required
I       that the calibration gases be contained in air, which is most
         likely to be the same carrier gas as that of the actual sample.
                                                          2
              Investigation of a reported oxygen synergism  has shown
I
•        that a 40/60 mixed fuel  (40 percent Hp»  60  percent  He)  is  required
          if the oxygen content of the emission  stream  varies more than  a
I        few percent from its  mean value.   Mixed  fuel  will also  be  re-
          quired if the oxygen  content of the emission  stream varies more
I        than a few percent from  the oxygen content  of the calibration  gases.
•             3.2  Organic compounds.   Acetylenic compounds  give a  slightly
          higher response than  aliphatic compounds.   Carbon atoms bound  to
•        oxygen, nitrogen, or  halogens give a reduced  or  zero  response.

I

I

-------
                               44
            3
     Table 1  illustrates these effects  in terms  of the  relative

response of one FIA to various  hydrocarbons.   The response  is

shown as effective carbon number (ECN),  as follows:


                          Instrument response  caused
                    ECN = by atom of given type
                          Instrument response  caused
                          by aliphatic carbon  atom

These values are true for one mode of operation of a specific

detector under specific conditions (e.g.,  mixed Np,  H« fuel).   It

has been reported that these numbers may vary  widely for different

operating conditions and for different detectors.  Variations  of

as much as 25 percent have been observed in studies of the  types

of organics associated with automotive emissions.  The variation

was observed to decrease with decreasing sample flow rate,  but

with an accompanying decrease in sensitivity.

         TABLE 1.  APPROXIMATE  EFFECTIVE CARBON NUMBERS
                   (FROM BECKMAN INSTRUMENTS)
Type of Atom
Carbon
Carbon
Carbon
Carbon
Carbon
Carbon
Oxygen
Occurrence
In Aliphatic Compound
In Aromatic Compound
In Olefinic Compound
In Acetylenic Compound
In Carbonyl Radical
In Nitrite
In Ether
Effective
Carbon Number
+1.0
+1.0
+0.95
+1.30
0.0
+0.3
-KO
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I

-------
I
I
I
                  TABLE 1.  APPROXIMATE EFFECTIVE CARBON NUMBERS
                                    CKMAN INSTRl
                                    (Continued)
I
I
•                           (FROM BECKMAN  INSTRUMENTS)

I

I

I

I

I

I
M            From this information it can  be seen that the accuracy of
         this  method for a given source will be largely dependent on the
•       particular makeup of organic emissions from the source.
              3.3  Other effects.  Significant changes in viscosity of the
|       emission gas from that of the calibration gas will affect the mass
_       rate  of organics to the detector.  If this phenomena is expected
•       to occur, a corrective technique must be devised.
•            If the instrument is calibrated with organic solvent standards,
         and then used to measure emissions of that solvent, their response
•       variations have been calibrated out.
         4.  Apparatus
Type of Atom
Oxygen
Oxygen
Oxygen
Chlorine

Chlorine
Nitrogen
Occurrence
In Primary Alcohol
In Secondary Alcohol
In Tertiary Alcohol, Ester
As two or more chlorine atoms
on single aliphatic carbon atom
In Olefinic Carbon Atom
In Amine
Effective
Carbon Number
-0.6
-0.75
-0.25

-0.12 each
+0.05
Value similar to
that for oxygen atom
in corresponding
alcohol
             4.1  Commercially available heated FIA.  The analyzer should

-------
                                                                         I
                                                                         I
be demonstrated, preferably by the manufacturer,  or his  repre-
sentative, to meet or exceed manufacturer's  specifications and            •
those described in this method.   The entire  sampling and analysis         •
system as encountered by gaseous organics must be capable of being
maintained in the temperature range of 350 to 440°F, or  less,             •
consistent with the emission regulation.
     4.2  Sample conditioning or interface system.  Probe with            I
filter, Teflon* sample line, Teflon-coated diaphragm pump or             •
stainless steel bellows pump and Teflon flow control valves,
capable of being maintained in the temperature range of  350  to            I
400°F, or less, consistent with the emission regulation.
     4.3  Potentiometric Recorder (optional).  Strip chart                |
recorder with a voltage output compatible with the FIA.                   m
5.  Reagents
     5.1  Fuel.  A hydrogen and helium mixture containing less            I
than 2 ppm organics (methane or carbon equivalent).
     5.2  Combustion Air.  High purity air with less than 2  ppm          |
organics  (methane or carbon equivalent).   Required only  if  the            .
emission  stream does not contain sufficient  oxygen.                      ™
     5.3  Zero Gas.  Less than 0.1 ppm organics (methane or               •
carbon equivalent).

* Mention of trade names on specific products does not constitute
endorsement by the Environmental Protection  Agency.
                                                                          I
                                                                          I
                                                                          I

-------
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
                 47
CL.
UJ
00
                                                           oo
                                                           O
                                                           QL
 L     ^
        d:
V
                                     O.
                                     f—
                                     O
                                     z:
                     O
                     OQ
                                                           O

                                                           2:
                                                           O
                                                    a.
                                                    UJ
                                                           CVJ

                                                           LU
                                                           a:
                                                           Z3
                                                           t!3

-------
                               48
I
I
I
     5.4  Calibration Gases  (2).   Gas  mixture  standards with  known
concentrations corresponding to  ranges of 5  to 10  ppm  and  5 to  10
percent (methane or carbon equivalent) are prepared  and certified
by a gas manufacturer.  The  mixture will  normally  consist  of                I
C-Hg  in air.  Other organic(s)  can be used,  if appropriate.  The
gas manufacturer must recommend  a  maximum shelf life for each              I
cylinder so that the concentration does not  change more than  +_ 5            •
percent from the certified value.   The date  of gas cylinder pre-
paration, certified propane  concentration and  recommended  maximum           I
shelf life must be affixed to the  cylinder before  shipment from
the gas manufacturer to the  buyer.  These gas  mixture  standards            |
are to be used to prepare a  calibration curve  as described in              •
Section 7.2.                                                               *
     5.5  Span Gas.  The calibration gas  corresponding to  5 to              I
10 percent (methane or carbon equivalent) is used  to span  the
analyzer.                                                                  |
     5.6  Organic Solvent.  Either a sample  obtained from  the              •
solvent source, or a sample  distilled from paint,  ink, etc. in
accordance with ASTM Procedure D3272-73T. Required  only if                 •
unaltered solvent emissions  are  being measured, mass calculations
in terms of the solvent are  necessary, and the relative response            |
factor of the FIA to the solvent is unknown.                               .
6.  System Performance Specifications                                      ™
     6.1  Linearity.  +_ 5 percent of the expected  value for full            •
scale settings up to the maximum percent absolute  (methane or
                                                                           I
                                                                           I
                                                                           I

-------
•                                      49
I
        carbon equivalent) calibration point.  The analyzer shall be
•      demonstrated prior to initial use to meet this specification
•      through a 5-point (minimum) calibration.  There shall be at
        least one calibration point in each of the following ranges:
•      5-10, 50-100, 500-1,000, 5,000-10,000, and 50,000-100,000 ppm
        (methane or carbon equivalent).  Certification of such demon-
•      stration by the manufacturer is acceptable.  An additional
•      linearity performance check (see Section 7.2.1) must be made
        before each use.
I           6.2  Zero Drift.  One percent full scale per test period.
             6.3  Span Drift.  One percent full scale per test period.
|      7.  Procedure
«           7.1  Sampling.
             7.1.1  Assemble the systems as shown in Figure 1.  Locate
•      the FIA in a suitably protected environment.  Take particular
        care that sample will be introduced to the FIA under the same
|      conditions of pressure and flow rates as are used in calibration.
_      For specific operating instructions for the FID, refer to
™      manufacturer's manual.
•           7.1.2  Adjust the sample conditioning and analyzer heating
        systems to provide a temperature of 350 to 400°F, or less, con-
•      sistent with the emission regulation, and allow the systems to
        warm up.
•           7.1.3  Perform a leak check as follows before sampling.
•      Recheck to confirm that all fittings are tight.  With the sample

I

I

-------
                               50
                                                                          I
probe plugged, open the  flow control  valve and the excess sample
bleed valve.   Use leak detection  fluid  or immerse the tubing               I
leading from  the bleed valve in a jar of water to check that               •
sample flow has ceased.   At the conclusion of the sampling tests,
recheck for leaks.                                                         I
     7.1.4  Begin Actual  Sampling.   Set the  signal attenuation
to yield a minimum response of 10 percent of full scale unless             |
the stack concentration  is  less than 1  ppm.  Adjust  the flow               .
and bleed valves to minimize sample  line residence time.  Com-
pare instrument readings with the calibration curve  to obtain              I
emission concentrations  based on  the calibration gas.
     7.1.5  At the conclusion of  the sampling tests,  but at                |
least once every day, introduce zero and span gases  to the                 _
analyzer to determine zero  and span  drifts.  If the  analyzer               "
has drifted beyond the allowable  performance specification,                •
the tests shall be considered invalid.
     7.2  Calibration and Solvent Standards.                              |
     7.2.1  Calibration  Curve. Maintain a  record of perform-
ance of each item.  Determine the linearity  of the analyzer as             •
follows:  With the signal attenuation at the most sensitive                •
setting, introduce zero  gas and adjust  the  respective zeroing
controls to indiate a reading of  less than  1 percent of full               I
scale.  With the signal  attenuation  at  the  least  sensitive
setting, introduce the span gas and  adjust  the span  control to             I
indicate the proper value on the  analyzer  readout.   Repeat  these
I
I
I

-------
 I
                                         51
 I
 _         two steps until  adjustments are no longer necessary.   Calculate  a
 "         predicted response for the 5 to 10 ppm calibration gas.   Introduce
 •         that calibration gas and note the value obtained.   If the value
           is not within +_ 5 percent of its predicted value,  then the
 I         analyzer may need repairs, or one or both of the calibration  gases
           may need replacement.  In any event, this linearity performance
 •         specification shall  be met before the analyzer is  placed  in
 •         actual use.
                7.2.2  Preparation of Solvent Standard Gas Mixtures.   (Optional--
 I         see Sections 1.2 and 5.6).  Assemble the apparatus shown  in Figure  2.
           Evacuate a 50-liter Tedlar or aluminized Mylar bag that has passed
 •         a leak check (described in Section 7.2.2.1) and meter in  about
 •         50 liters of air.  Measure the barometric pressure, the relative
           pressure at the  dry gas meter, and the temperature at the dry gas
 •         meter.  Mhile the bag is filling use the 10 yl  syringe to inject
           10 yl  of the solvent through the septum on top of  the impinger.
 I         This gives a concentration of approximately 200 yg/liter.
 •         In a like manner, use the other syringe to prepare dilutions  having
           approximately 40 and 20 yg/liter concentrations.   To  calculate the
 •         specific concentrations, refer to Section 8.1.   These gas mixture
           standards may be used for a few days from the date of preparation,
 |         as determined by repetitive analysis for concentration degradation.
M         (Caution:  Contamination may be a problem when a bag  is reused if the
           new gas mixture  standard is a lower concentration  than the previous
I         gas mixture standard.)

I

I

-------
                                 52
I
I
     7.2.2.1   Solvent Standards  Bag  Leak Checks.   While  performance
of this section is required subsequent to bag  use, it  is also advised        I
that it be performed prior to bag use.  After  each use,  make sure
a bag did not develop leaks as follows:   to  leak  check,  connect a            I
water manometer and pressurize the bag to 5-10 cm H20  (2-4  in. hLO).         •
Allow to stand for 10 minutes.  Any  displacement  in the  water mano-
meter indicates a leak.   Also, check the rigid container for leaks           I
in this manner.  (Note:   an alternative leak check method  is to
pressurize the bag to 5-10 cm HLO or 2-4 in. HLO  and allow  to stand          |
overnight.  A deflated bag indicates a leak.)   For each  sample bag           »
in its rigid container,  place a  rotameter in line between  the bag
and the pump inlet.  Evacuate the bag.  Failure of the rotameter to          •
register zero flow when  the bag  appears to be  empty indicates a leak.
8.  Calculations                                                            |
     All measurements or calculations must be  corrected  for CH., if          _
required by the emission regulation.                                        "
     8.1  Carbon or Surrogate Organic Compound Concentration.                •
     8.1.1  Volume concentration [ppm].  To  determine  emission concen-
trations of total gaseous organics (wet basis) on a CH.  or carbon            |
equivalent basis, multiply the recorded emission  values  by the number        _
of carbon atoms in a molecule of calibration gas.  In  some  instances         •
it will be required to report emissions on the basis of  the calibra-         •
tion gas, in which case no calculations are  necessary.
     8.1.2  Mass concentration [mg/m ].  To  convert volume concentra-        |
tion to mass concentration, proceed as follows:                             _
                                                                            I
                                                                            I

-------
I

I
                                       53
I

I
              8.1.2.1  Establish Standard Conditions.  Find the volume
|        occupied  by  1 mg. mole of ideal gas at these conditions.  Then
                                            2
•        find the  number of mg. moles in 1 m   (at saturation).
              8.1.2.2 Determine the molecular weight of the assumed organic
I        compound  in  which the emission is to be expressed.
              8.1.2.3  Use the values obtained in 8.1.2.1 and 8.1.2.2 to

          determine the mass concentration  (mg/m ) at saturation.  Divide this
                     C                 •}
          number by 10  to find the mg/m  equivalent to 1 ppm.
              8.1.2.4  Multiply the result obtained in 8.1.2.3 by the volume

          concentration obtained in 8.1.1.  The result is the mass concen-

          tration  expressed  in terms of the compound whose molecular weight

•        was determined  in  8.1.2.2.

              8.2 Organic  Solvent Concentration

•            8.2.1  Solvent Standards Concentrations.  Calculate each solvent

•        standard concentration prepared in accordance with Section 7.2.1.2

          as follows:                               3
                                           I. mo) 10	y_q
I                                 r  =      ul    mg
                                   c  = 	Ml    "'tf            Equation 1
                                    c  ..  u 293  m
         where:
              C  = Solvent standard concentration, yg/1.

              B  = Number of yl of solvent injected.
I

•            Y  = Dry gas meter, calibration factor.

I
•            d. = Density of the solvent at 293°A.


I
              V  = Gas volume measured by dry gas meter in liters.
              P  = Absolute pressure of the dry gas meter, mm Hg.

              T  = Absolute temperature of the dry gas meter, °A.

-------
                                                                           I
                                 54
     8.2.2  Solvent Emission  Concentrations  The emission values            I
in pg/ml  are taken from  the solvent  standards response curve.  Mo           •
further calculations are required.
                                                                           I
9.  References
     1.  Method 108, 43101-02-71T,  "Tentative Method  for  Continuous
Analysis of Total Hydrocarbons  in the Atmosphere  (Flame  lonization          |
Method)."  Methods of Air Sampling and Analysis,  Intersociety               _
Committee, Amer. Pub. Health Assn., Washington, D.  C., 1972.                •
     2.  M. Johnson, "Oxygen Synergism in the Model 400  FIA."               •
Beckman Instruments, Inc., Fullerton, Ca. Oct., 1970.
     3.  Instruction Manual  82132-A, "Model  402 Hydrocarbon                 |
Analyzer."  Beckman Instruments, Inc., Fullerton, Ca.  Feb.,  1971.
     4.  "Air-Hydrocarbon Monitoring Instrumentation," Lawrence            •
Berkeley Laboratory, Univ. of Ca., Berkeley, Ca.  Nov., 1973.                •
     5.  A. J. Andreatch and R.  Feinland, "Continuous Trace  Hydro-
carbon Analysis by Flame lonization." Anal.  Chem. 32  (8)  1021-4.            •
July, 1960.
     6.  R. A. Morris and R. L.  Chapman, "Flame  lonization Hydro-           •
carbon Analyzer." J. Air. Pol.  Cont. Assoc.  11 (10) 467-9. Oct.,  1961.      •
     7.  F. M. Black, L. E.  High, and J. E.  Sigsby, "The Application
of Total Hydrocarbon Flame lonization Detectors  to the Analysis of          I
Hydrocarbon Mixtures from Motor Vehicles, With and Without Catalytic
Emission Control." Water. Air. Soil Pollut.  5 (1) 53-62.  Oct., 1975.       I

                                                                           I

                                                                           I

                                                                           I

                                                                            I

-------
1
m TECHNICAL REPORT DATA
(Please read Instructions on the "everse bcjore completing)
j. F*>ORT NO. 2.
Ep-450/2-78-041
t. TITLE ANDSUBTITLE
MBisurement of Volatile Organic Compounds
1. AUTHOR(S)
Epssion Measurement Branch
). PERFORMING ORGANIZATION NAME AND ADDRESS
EHjssion Measurement Branch (MD-13)
Ejironmental Standards and Engineering Division
U. S. Environmental Protection Agency
Rwearch Triangle Park, North Carolina 27711
I2.HONSORING AGENCY NAME AND ADDRESS
Dm for Air Quality Planning and Standards (MD-10)
Office of Air, Noise, and Radiation
U«S. Environmental Protection Agency
Rmearch Triangle Park, North Carolina 27711
3. RECIPIENT'S ACCESSION NO.
5. REPORT DATE
October, 1978
6. PERFORMING ORGANIZATION CODE
8. PERFORMING ORGANIZATION REPORT NO.
10. PROGRAM ELEMENT NO.
11. CONTRACT/GRANT NO.
13. TYPE OF REPORT AND PERIOD COVERED
14. SPONSORING AGENCY CODE
EPA/200/04
15. SUPPLEMENTARY NOTES
1
   BSTRACT

I      This document discusses the rationale of total volatile organics stationary
    rce emission measurement through the determination of organic carbon mass
 concentration.   A conceptual approach for writing emission regulations in terms
 o^volatile organic carbon is recommended, and drafts of two specific test methods
 aB presented for regulation implementation.  The methods are the measurement of
 tRal gaseous nonmethane  organics as carbon by the chromatographic oxidation/
 reduction procedure, and the relative organic measurement derived by direct appli-
 cBion of the flame ionization analyzer.
  I

  I
I
7. KEY WORDS AND DOCUMENT ANALYSIS
DESCRIPTORS
Mr Pollution
Analyzing
Imp! ing
ganic Compounds
Gas Sampling
8. IK I HIBUTION STATEMENT
•lease Unlimited
b. IDENTIFIERS/OPEN ENDED TERMS
Stationary Sources
Volatile Organic
Compounds
Analytical Strategy
Organic Vapors
Environmental Assessment
19. SECURITY CLASS (This Report)
Unclassified
20. SECURITY CLASS (This page)
Unclassified
c. COS AT I Field/Group
13B
21. NO. OF PAGES
54
22. PRICE
:PA Form 2220-1 (Rev. 4-77)   PREVIOUS EDITION is OBSOLETE
  I

-------
                                                                                                                               5 co  n
                                                                                                                                
 .
^ —   CD

O Ctl    ^J"
                                                                                                                     m ^ TJ m TI ~v
                                                                                                                     ?"§ a ? 8 g
                                                                                                                           3   Q-
                                                                                                                                            ?! -n"

-------