United States         Air and Radiation       EPA340/1-91-011
EPA/340/1-91/011          nental Protection    (EN-341W)         September 1991
                 mdard Operating Procedure
              for Analysis of Ink ^
              Samples by Reference
              Methods 24 and 24A
                                          1445 ROSS AVENUE
                                             , TEXAS 7520?

-------
                                  EPA-340/1-91-011
 STANDARD OPERATING PROCEDURE FOR
 ANALYSIS OF COATING AND INK SAMPLES
   BY REFERENCE METHODS 24 AND 24A
U.S. ENVIRONMENTAL PROTECTION AGENCY
        Office of Air and Radiation
 Office of Air Quality Planning and Standards
   Stationary Source Compliance Division
          Washington, DC 20460

            September 1991

-------

-------
                                 DISCLAIMER
      This document is a draft report submitted to the Stationary Source Compliance
Division for circulation and subsequent comment on the technical adequacy of the
contents. Any view or opinions contained herein are those of the authors and do not
necessarily reflect the conclusions of the U.S. Environmental Protection Agency.

      Any mention of trade names or commercial products does not constitute
endorsement or recommendation for use by the U.S. Environmental Protection
Agency.

-------

-------
                                CONTENTS



Section                                                              Page

1.     INTRODUCTION  	1

2.     SAMPLE HANDUNG PROCEDURES	3

3.     REFERENCE METHOD 24	7

4.     SOP FOR METHOD 24 ANALYSES	15

           Determination of Percent Water Content in Coating Samples
           by Gas Chromatography  	19

           Determination of Water Content in Coating by Karl Fischer
           Titration	31

           Determination of Percent Volatile Content  in Coating
           Samples	39

           Determination of Density of Paint, Varnish, Lacquer, and
           Related Products	43
                                                      •
           Determination of Dichloromethane and 1,1,1-Trichloroethane
           in Paints and Coatings by Direct Injection  into a Gas
           Chromatograph	47

5.     REFERENCE METHOD 24A	63

6.     SOP FOR METHOD 24A ANALYSES	67

7.     BIBLIOGRAPHY	69
                                                                    Page

-------
                                  SECTION 1

                                INTRODUCTION
      This standard operating procedure document is prepared with the intent of
providing simple step-by-step instructions, covering all aspects of laboratory handling
and analysis of coating and ink samples, for use by EPA, State, and local regulatory
agency laboratories as well as contract laboratories nationwide. The instructions are
presented in general terms as much as possible while providing sufficient details for
accurate and precise laboratory measurements to be done. The procedures should
be used in conjunction with existing health and safety programs and in accordance
with existing EPA or other agency training guidelines.

      Standard procedures are presented in this report for the handling  and
physical/chemical characterization of coating and ink samples that require analysis by
EPA Reference Methods 24 or 24A (RM 24/24A) as found in 40 CFR 60, Appendix A.
This document has been  expanded from an original version obtained from the
Wisconsin Occupational Health Department covering laboratory determinations of the
VOC content, water content, and density of coating and ink samples taken for
environmental compliance.  For the purpose of developing consistency among
analytical laboratories  nationwide, it delineates  the activities deemed proper and
necessary to ensure that  an accurate and precise measurement of coating or ink
sample properties are made.  If the activities are conducted as described, questions
concerning the validity and reproducibility of analyses performed can be avoided.

      The purpose of these standard operating procedures (SOPs) for RM 24 and RM
24A is to ensure that laboratory analyses are performed in such a way that review,
interpretation, and use of the laboratory analysis results by EPA, State, and local
regulatory agency personnel is facilitated.

      The procedures cover all types of industrial coating and printing ink operations,
regardless of the method of application of coating or ink,  including, for example, dip,
spray, roll,  flow, electrostatic, or electro-deposition processes.  Special cases involving
use of exempt solvents dichloromethane  (methylene chloride)  and 1,1,1-trichloroethane
(methyl chloroform) or multi-component mixture coating formulations are  covered in
accordance with most recent technical and regulatory thinking, however,  this is not a
regulatory but a technical procedure guideline  document.

      In addition to the introduction given here, Section 2 of this document contains a
summary of sample handling procedures to be followed; Section 3 contains  EPA
Reference Method 24  including draft revision amendments which are being considered
to make it applicable to multi-component coatings and exempt solvents; Section 4
contains detailed SOPs for EPA Method 24 analyses for percent water content,


                                                                        Page 1

-------
percent volatile content, exempt solvent (dichloromethane and 1,1,1-trichloroethane),
and density of coatings; Section 5 contains EPA Reference Method 24A for
determination of volatile matter and density of printing inks; Section 6 contains SOP
guidance for EPA Method 24A analyses; and Section 7 contains a detailed
bibliography of documents relevant to EPA Method 24/24A analyses.
Page 2

-------
                                 SECTION 2

                      SAMPLE HANDLING PROCEDURES
      There are several steps to be taken at the laboratory upon receipt of a sample
for analysis according to EPA Reference Methods 24 and 24A.  These are outlined
below:

      1.     Inspect the shipping package for damage and proper labeling.  If
            improper labeling, delayed shipping, damage, spillage, or other problem
            is evident, notify the sending agency by phone as soon as possible.

      2.     Open the shipping package and record in a laboratory notebook or
            recordbook relevant observations and information from containers,
            labels, shipping papers, Material Safety Data Sheets (MSDS) forms, and
            Chain of Custody (C of C) documents, which should be enclosed with
            the samples.  Include  the following:

            2.1    Name of person who collected the samples, the date/time the
                  samples were obtained,  and the date/time the samples were
                  shipped to the laboratory for analysis.

            2.2   Name, phone number, and address of agency and person to
                  whom results of analyses are reported.

            2.3   Facility or source identification to be coded or numbered.

            2.4   Sample identification number and a brief description of the sample
                  and container.

            2.5   Analyses required, RM 24, RM 24A, multi-component, or exempt
                  solvent determinations.

            2.6   Describe the physical appearance of the sample on first opening
                  (e.g., full to top, not tightly sealed or inner seal missing,
                  appearance (color), type of ink/coating, odor, or presence of
                  residue).

            2.7   Summarize MSDS information, if available, for safety and handling
                  purposes.
                                                                       Page 3

-------
            2.8   If special multi-component sample blending or exempt solvent
                  determination is required, contact sending agency personnel for
                  instructions on set-up (hardening) time and mixing proportions.

            2.9   Report problems with sampling integrity or inconsistencies in
                  documentation, including missing (or occasionally too much)
                  information on the labels or C of C forms.

            2.10  Start any internal laboratory sample recordkeeping procedures
                  normally used to track samples.

      3.     After signing the C of C receipt of the samples,  send the original of the
            C of C documents back to the  appropriate sending agency personnel for
            their records and keep a copy for laboratory records.

      4.     Secure, the samples with a copy of the C of C documents and MSDS
            forms in a locked storage cabinet dedicated to environmental samples.  If
            this is done by a sample custodian designated to retain custody of the
            sampling pending analysis, then that person should  sign the C of C
            forms acknowledging receipt of the samples.  Maintain samples  at room
            temperature, preferably at 70°F but within the range  of 40 to 100°F.

      During and at the end of sample analyses, there  are several steps to be taken
related to disposal of waste materials from the sample and the laboratory analysis.
These are summarized  below:

      5.     A plastic bag should be used to discard waste ink solids, rags, paper
            towels, kimwipes, and other non-glass or non-sharp items.  This can
            normally be disposed of with industrial solid waste material.

      6.     A separate container (e.g., a glass bottle or metal/plastic can), should be
            used for collection and eventual disposal of discarded glassware and
            other sharp-edged items used in the laboratory.  This may be discarded
            with industrial solid waste, but must be given  careful handling during
            disposal.

      7.     A 5-gallon solvent-resistant plastic or metal container with a lid or cover
            designed for flammable liquids  should be stored in the hood where
            analytical work is performed.  It should be used to collect unused or
            excess sample, reagents, and mixtures of  both for periodic disposal as a
            liquid waste.  Fiber mesh material can be used to filter out solids from
            liquid wastes that are deposited into  the container.  A list of compounds
            which are placed in the container should be kept as a record for
Page 4

-------
      identification of the contents upon disposal. Liquid waste should be sent
      out for disposal to a reputable disposal organization.

8.     After analyses are completed (or approximately 30 days), the samples in
      their sample containers should be attached to their MSDS forms identi-
      fying the approximate composition, physical and chemical properties, and
      safety hazards associated with the sample.  They should then be sent
      out for disposal to a reputable disposal organization.  If analyses have
      not been completed within 30 days,  it is recommended that another
      sample should be taken since the chemical/physical composition of the
      existing sample may have changed significantly over the time period.
                                                                  Page 5

-------
Page 6

-------
                                 SECTION 3

      REFERENCE METHOD 24-DETERMINATION OF VOLATILE MATTER
    CONTENT AND DENSITY OF PRINTING INKS AND RELATED COATINGS
      EPA Reference Method 24 as it appears in 40 CFR Part 60, Appendix A (7-1-90
edition) is presented here, starting on the following page, with draft revisions under
consideration by EPA shown in bold italics. These revisions incorporate procedures to
be used with multi-component coatings and exempt solvents which are not addressed
in the 7-1-90 edition of 40 CFR 60, Appendix A.

      Analysis for exempt solvents in coatings is practiced according to ASTM D4457-
85 - "Standard Test Method for Determination of Dichloromethane and 1,1,1-Trichloro-
ethane in Paints and Coatings by Direct Injection into a Gas Chromatograph."  In the
draft revisions to EPA Method 24, this procedure is incorporated by reference.

      Proposed guidance for handling multi-component coatings, developed for
incorporation into EPA Method 24, includes revisions to the existing text and some
additional text. In the revisions, existing Sections 3.1, 3.2, 3.3, and 3.4 have been
redesignated as Sections 3.2, 3.3, 3.4, and 3.5, respectively; Sections 5.1 and  5.2
have been revised; and new Sections 2.5, 3.1, 3.6, and 3.7 have been added,  as
shown in bold italics on the following  pages.
                                                                       Page 7

-------
REFERENCE METHOD 24 - DETERMINATION OF VOLATILE MATTER CONTENT,
WATER CONTENT, DENSITY, VOLUME SOLIDS, AND WEIGHT SOUS OF SURFACE
COATINGS
                                     Current Draft Revisions Shown in Bold Italics
40 CFR Part 60, Appendix A
(7-1-90 Edition)
Final, promulgated 10/3/80
45 FR 65958
1.     Applicability and Principle

      1.1    Applicability. This method applies to the determination of volatile matter
            content, water content, density, volume solids, and weight solids of paint,
            varnish, lacquer, or related surface coatings.

      •1.2    Principle. Standard methods are used to determine the volatile matter
            content, water content, density, volume solids, and weight solids of paint,
            varnish, lacquer, or related surface coatings.

2.     Applicable Standard Methods

      Use the apparatus, reagents, and procedures specified in the standard
methods below:

      2.1    ASTM D1475-60 (Reapproved 1980), Standard Test Method for Density
            of Paint, Varnish, Lacquer, and Related Products (incorporated by
            reference-see §60.17).

      2.2    ASTM D2369-81, Standard Test Method for Volatile Content of Coatings
            (incorporated by reference-see §60.17).

      2.3    ASTM D3792-79, Standard Test Method for Water Content of Water-
            Reducible Paints by Direct Injection into  a Gas Chromatograph (incorpo-
            rated by reference-see §60.17).

      2.4    ASTM D4017-81, Standard Test Method for Water in Paints and Paint
            Materials by the Karl Fischer Titration Method (incorporated by reference-
            -see§60.17).

      2.5    ASTM D4457-35 Standard Test Method for Determination of Dichloro-
            methane and 1,1,1-Trichloromethane in Paints and Coatings by Direct
            Injection into a Gas Chromatography.
                                                                       Page 9

-------
3.     Procedure
      3.1   Multi-Component Coatings.  To determine the total volatile content, water
            content, and density of multi-component coatings, follow the procedures
            in Section 3.7.  Multi-component coatings are coatings that are packaged
            in two or more parts, which are combined before application.  Upon
            combination a coreactant from one part of the coating chemically reacts,
            at ambient conditions, with a coreactant from another part of the coating.
            For all other coatings analyze as follows:

      3.2   Volatile Matter Content.  Use the procedure in ASTM D2369-81
            (incorporated by reference-see §60.17) to determine the volatile matter
            content (may include water) of the coating. Record the following
            information:

                  W,  =  Weight of dish and sample before heating, g.
                  W2 =  Weight of dish and sample after heating, g.
                  W3 =  Sample weight, g.

            Run analyses in pairs (duplicate sets) for each coating  until the criterion
            in Section 4.3 is met.  Calculate the weight fraction of the volatile matter
            (WJ for each analysis as follows:

                                       W -W
                                   W--^^                          Eq. 24-1
            Record the arithmetic average  (WJ.


      3.3   Water Content. For waterborne (water reducible) coatings only,
            determine the weight fraction of water (WJ using either "Standard
            Content Method Test for Water of Water-Reducible Paints by Direct Injec-
            tion into a Gas Chromatograph" or "Standard Test Method for Water in
            Paint and Paint materials by Karl Fischer Method." (These two methods
            are incorporated by reference - see §60.17.)  A waterborne coating is
            any coating which contains more than 5  percent water by weight in its
            volatile fraction.  Run duplicate sets of determinations until the criterion in
            Section 4.3 is met.
            Record the arithmetic average  (WJ.
Page 10

-------
3.4   Coating Density. Determine the density (Dc, kg/liter) of the surface
      coating using the procedure in ASTM D1475-60 (Reapproved 1980)
      (incorporated by reference - see §60.17).  Run duplicate sets of
      determinations for each coating until the criterion in Section 4.3 is met.

      Record the arithmetic average (D£.


3.5   Solids Content. Determine the volume fraction (VJ solids of the coating
      by calculation using the manufacturer's formulation.

3.6   Exempt Solvent Content Determine the weight content of exempt
      solvents (WJ by using ASTM Method D4457-85 (incorporated by
      reference - see §60.17). Run a duplicate set of determinations.
      Record the arithmetic average

      Note: exempt solvents are defined as those solvents listed in 57 FR
      3941, February 3, 1992.  Dichloromethane and 1,1,1-trichloroethane are
      listed exempt solvents and may be used in coatings.

3.7   To determine the total volatile content, water content, and density of
      multi-component coatings, use the following procedures:

      3.7.1  Prepare about 100 ml of sample by mixing the components in a
            storage container, such as a glass jar with a screw top or a metal
            can with a cap. The storage container should be just large
            enough to hold the mixture. Combine the components (by weight
            or volume) in the ratio recommended by the manufacturer. Tightly
            close the container between additions and during mixing to
            prevent loss of volatile materials.  However, most manufacturers
            mixing instructions are by volume. Because of possible error
            caused by expansion of the liquid when measuring the volume, it
            is recommended that the components be combined by weight
            When weight is used to combine the components and the
            manufacturer's recommended ratio is by volume, the density must
            be determined by Section 3.4.

      3.7.2  Immediately after mixing, take aliquots from this 100 ml sample for
            determination of the total volatile content, water content, volume
            solids, and density.  To determine water content, follow Section
            3.3.  To determine density, follow Section 3.4.  To determine
            volume solids content, follow Section 3.5.  To determine total
            volatile content, use the apparatus and reagents described in
                                                                 Page 11

-------
                  ASTM D2369-81, Sections 3 and 4, respectively, and the following
                  procedures:

                  3.7.2.1      Weigh and record the weight of an aluminum foil
                              weighing dish. Add 3  ± 1 ml of suitable solvent as
                              specified in ASTM D2369-81 to the weighing dish.
                              Using a syringe as specified in ASTM D2369-81,
                              weigh to 1 mg, by difference, a sample of coating
                              into the weighing dish.  For coalings believed to
                              have a volatile content less than 40 weight percent, a
                              suitable size is 0.3 ± 0.10 g, but for coatings
                              believed to have a volatile content greater than 40
                              weight percent a suitable size is 0.5 ± 0.10 g. Note:
                              If the volatile content determined pursuant to Section
                              5 is not in the range corresponding to the sample
                              size chosen repeat the test with the appropriate
                              sample size. Add the specimen dropwise, shaking
                              (swirling) the dish to disperse the specimen
                              completely in the solvent  If the material forms a
                              lump that cannot be dispersed, discard the specimen
                              and prepare a new one.  Similarly, prepare a
                              duplicate. The sample shall stand for 1 hour, but no
                              more than 24 hours prior to being oven dried at
                              110°C  ± 5°C for 1 hour.

                  3.7.2.2      Heat the aluminum foil dishes containing the
                              dispersed specimens in the forced draft oven for 60
                              minutes at 110 ±5°C.  Caution - provide  adequate
                              ventilation, consistent with accepted laboratory
                              practice, to prevent solvent vapors from accu-
                              mulating to a dangerous level.

                  3.7.2.3      Remove the dishes from the oven, place immediately
                              in a desiccator, cool to ambient temperature, and
                              weight to within  1 mg.

                  3.7.2.4      Run analyses in pairs (duplicate sets) for each
                              coating mixture until the criterion in Section 4.3 is
                              met Calculate W, following Equation 24-1 and
                              record the arithmetic average.
Page 12

-------
4.
Data Validation Procedures
      4.1
      4.2
      Summary.  Trie variety of coatings that may be subject to analysis makes
      it necessary to verify the ability of the analyst and the analytical
      procedures to obtain reproducible results for the coatings tested. This is
      done by running duplicate analyses on each sample tested and
      comparing results with the within-laboratory precision statements for
      each parameter. Because of the inherent increased imprecision in the
      determination of the VOC content of waterborne coatings as the weight
      percent water increases, measured parameters for waterborne  coatings
      are modified by the appropriate confidence limits based on between-
      laboratory precision statements.

      Analytical  Precision Statements. The within-laboratory and between-
      laboratory precision statements are given below

Volatile Matter
Content, Wv

Water Content, Ww
Density, Dc
Within-laboratory
1.5% W7

2.9% Ww
0.001 kg/liter
Between-laboratory
4.7 % W7

7.5% Ww
0.002 kg/liter
      4.3
      4.4
      Sample Analysis Criteria.  For Wv and Ww, run duplicate analyses until the
      difference between the two values in a set is less than or equal to the
      within-laboratory precision statement for that parameter.  For Dc run
      duplicate analyses until each value in a set deviates from the mean of the
      set by  no more than the within-laboratory precision statement.  After
      several attempts, if it is concluded that the ASTM procedures cannot be
      used for the specific coating with the  established within-laboratory
      precision, the Administrator will assume responsibility for providing the
      necessary procedures for revising the method or precision statements
      upon written request to:  Director, Emission Standards and Engineering
      Division, (MD-13) Office of Air Quality Planning and Standards, U.S.
      Environmental Protection Agency, Research Triangle Park,  NC 27711.

      Confidence Limit Calculations for Waterborne Coatings. Based on the
      between-laboratory precision statements, calculate confidence limits for
      waterborne coatings as follows:
                                                                        Page 13

-------
                  To calculate the lower confidence limit, subtract the appropriate
                  between-laboratory precision value from the measured mean value
                  for that parameter.

                  To calculate the upper confidence limit, add the appropriate
                  between-laboratory precision value to the measured mean value
                  for that parameter.

                  For Wv and Dc, use the lower confidence limits, and for Ww, use
                  the upper confidence limit. Because Vs is calculated, there is no
                  adjustment for the parameter.
5.     Calculations

      5.1  Nonaqueous Volatile Matter.

            5.1.1  Solvent-borne Coatings.
                                    W0*WV                           Eq. 24-2
            Where:

            W0  = Weight fraction nonaqueous volatile matter, g/g.

            5.1.2 Waterborne Coatings.
                                  Wa=Wv-Ww                         Eq. 24-3
            5.7.3 Coatings Containing Exempt Solvents.
      5.2 Weight Fraction Solids.


                                   W3*-\-Wv                          Eq.24-5


            Where:

            Ws  =  Weight solids, g/g.
Page 14

-------
                                 SECTION 4

      STANDARD OPERATING PROCEDURES FOR METHOD 24 ANALYSES


PRIMARY ASTM METHODS

      EPA Method 24 calls for the determination of density by ASTM D1475-60
(Reapproved 1980), volatile content by ASTM D2369-S1, and water content by either
ASTM D3792-79 (gas chromatography) or ASTM D4017-81  (Karl Rscher titration) of
coating samples taken by regulatory agency personnel.  ASTM D4457-85 is
considered the best guidance presently available to determine the exempt solvent
content of coatings.


SAMPLE CHARACTERIZATION STEPS

General

      It is essential to make use of all the characterization information that may be
available  on the sample to proceed expeditiously.   If the MSDS forms are available with
the samples (as they should be), it can be determined if a highly volatile solvent, such
as a light petroleum ether, is present or if the sample consists of more than 70-75%
volatile solvents.  Otherwise, a phone call can be made to the sending agency to
determine the approximate solvent composition expected.  This helps to determine the
amount of sample to be used in the volatile content analysis to obtain a good reading
on the volatile content, using ASTM D2369-81.
                                                     •

Step 1

      The first step to be taken with a sample is opening and examining it in the
hood. Important points to note here include amount of sample relative to container
size (full,  almost full, 3/4 full, etc.), color, viscosity,  and uniformity.  If questions arise at
this point because of a conflict of observation with  included information, the sending
agency must be contacted  to determine the proper course of action.

Step 2

      If specific exempt solvent assay is requested, then the preparation of aliquots of
sample for the determination of the content of the solvent should begin right after the
sample examination. The standard test method ASTM D4457-85 for determination of
dichloromethane (methylene chloride) or 1,1,1-trichloroethane (methyl chloroform) is
being considered for incorporation into EPA Method 24. If the exempt solvent assay  is
not run immediately these aliquots should be securely sealed and stored at -20 C until


                                                                       Page 15

-------
the assay is performed. At any rate, during the performance of analytical procedures,
ideally no interruptions should be allowed and sample containers should always be
kept sealed except when they must be opened to take or transfer sample material.
The results of assays for exempt solvents (and of the water assay to be done later in
the procedure) will be subtracted from the total volatile content.

Step 3

      The total volatile content  will be determined using ASTM D2369-81.  This should
follow the previous step with minimal delay  to ensure the best estimate of the total
volatile content.

Step 4

      The next step in the procedure should be the analysis for water content by
either ASTM  D3792-79 (gas chromatography) or ASTM D4017-81 (Karl Rscher
titration). One of two conditions should exist before this step is included:  either the
sending agency specifically requests analysis for water content or the documentation
accompanying the sample indicates the presence of water.  Care is taken  during all
previous steps to prevent the coating sample from absorbing water.  The total  volatile
content reported would be too high if the sample absorbed water in the lab.

Step 5

      The density determination using ASTM D1475-60 (Reapproved 1980) is always
done last because it entails use of a large quantity of sample and would provide the
maximum opportunity for sample volatilization.
SOPs FOR REFERENCE METHOD 24 ANALYSES

      The following five procedures have been adapted from the pertinent ASTM
methods to provide additional detail as well as to provide step-by-step procedures to
be followed in their use under EPA Method 24.  The five procecdures are presented in
the following order:

      •     Determination of Percent Water Content in Coating Samples by Gas
            Chromatography (Adapted from ASTM D3792-79)

      •     Determination of Water Content in Coatings by Karl Fischer Titration
            (Adapted from ASTM D4017-81)

            Determination of Percent Volatile Content  in Coating Samples (Adapted
            from ASTM D2369-81)


Page 16

-------
Determination of Density of Paint, Varnish, Lacquer, and Related
Products (Adapted from ASTM D1475-60)

Determination of Dichloromethan and 1,1,1-Trichloroethane in Paints and
Coatings by Direct Injection into a Gas Chromatograph (Adapted from
ASTM D4457-85)
                                                          Page 17

-------
Page 18

-------
DETERMINATION OF PERCENT WATER CONTENT IN COATING SAMPLES BY GAS
CHROMATOGRAPHY (Adapted from ASTM D3792-79)
METHOD:   The percent water content is determined by direct injection
            into a gas chromatograph.

EQUIPMENT:

I.     Glassware

      A.     A bottle of 50 ml capacity with a cap having an inert liner - one.

      B.     20 ml scintillation vials with teflon inverted cone cap seals  - three.

      C.     10 ml serum bottles - six + four per sample.

            1.    Teflon coated septa - six + four per  sample.

            2.    Aluminum caps - six + four per sample.

II.     Syringes and pipettes

      A.     10 ml disposable syringe with needle - one.

      B.     5  ml disposable syringe with needle - two.
              •
      C.     A 5 ml pipetting device with one disposable tip.

      D.     A 200 n\ Eppendorf pipettor with three disposable tips.

      E.     Transfer pipette (5 3/4" Pasteur type) - five  per sample.

      F.     A 10 /*! syringe with needle.

III.    Chemicals

      A.     Deionized water (H2O).

      B.     Anhydrous N,N-Dimethylformamide (DMF).

      C.     Anhydrous 2-propanol (i-PrOH).

      D.     Anhydrous n-propanol (n-PrOH).
                                                                    Page 19

-------
      E.    High purity, dry nitrogen (NJ.

IV.    Miscellaneous

      A.    Rubber tubing - 18 inches.

      B.    Bottle cap crimper.

      C.    Pressure regulator for N2 tank.

V.    Analytical balance - Range 0-200 grams, Reproducibility: ±0.1 mg

VI.    Gas Chromatograph  (GC) - e.g., HP 5890 A

      A.    Thermal conductivity detector.

      B.    Porapak Q column -  10 ft.

VII.   Integrator - e.g.,  HP 3357 LAS or HP 3392 A.

      Note: Anhydrous solvents in packaging similar to the Sure/Seal
            Packaging System employed by the Aldrich Chemical Company,
            Inc. allow  storing and dispensing of the solvent without exposure
            to atmospheric moisture. This procedure describes using
            reagents packaged in that manner, but the steps described  can
            cause deterioration of the septa and exposure of the solvent to
            atmospheric moisture. An alternative procedure would be used in
            a high  humidity environment or to ensure the transfer of solvents
            as free of moisture as possible.

PROCEDURE:

I.     General Preparation

      A.    Assemble equipment needed.

      B.    Fill a 20 ml scintillation vial with deionized water, cap it, label the vial
            "H2O."

      C.    Label a 10 ml syringe with needle and the 50 ml bottle "DMF".  One 5 ml
            syringe with needle and  a 20 ml scintillation vial should be labeled "i-
            PrOH"  and another 5 ml syringe with needle and 20 ml scintillation vial
            should be labels "n-PrOH".


Page 20

-------
D.    The bottles containing anhydrous DMF, anhydrous i-PrOH and
      anhydrous n-PrOH can be slightly pressurized relative to the surrounding
      atmosphere with high purity, dry N2 to facilitate the withdrawal of material.

      1.     Attach rubber tubing to N2 supply manifold and close the other
            outlets.

      2.     Turn on the N2 to a very gentle flow (at 3-5 psig).

      3.     Pressurize the DMF bottle while wearing gloves, remove the
            needle from the DMF syringe and insert the needle base in the
            open end of the tubing. (TAKE GREAT CARE to hold the needle
            in the tube firmly or it will be released at this point like a projectile.)

      4.     Insert the needle in the septum of the DMF bottle and invert the
            bottle.

      5.     When the bubbles decrease, right the bottle and remove the
            needle.

      6.     Replace the needle on the DMF syringe.

      7.     Repeat the procedure for i-PrOH and n-PrOH using the
            appropriate needle if needed.

      8.     Turn off the N2 flow.

E.    Using the 10 ml syringe labelled  "DMF," withdraw the approximate
      amount of DMF needed (2 ml per serum bottle used  in screening and
      analysis). Transfer the DMF to the labeled 50 ml bottle and cap.

Preparation of Screening Samples

A.    Label the 10 ml serum bottles needed for this step.  One should be
      labeled "DMF," another "DMF + i-PrOH + n-PrOH." Additional bottles,
      one for each sample, should  be labeled using the appropriate laboratory
      sample number.

B.    Pipet 2 ml DMF into the bottle labelled "DMF."  Place septum (teflon side
      down) on top of the  bottle, cover both with metal cap and crimp.

      1.     The action of the crimping tool should be  gentle but firm.

      2.     The cap should not move if sealed properly.


                                                                Page 21

-------
            3.     If it is loose, rotate the vial 1 /4 turn and crimp again.

            4.     Repeat until tightly sealed.

      C.    Pipet 2 ml DMF, 200 n\ i-PrOH and 200 //I n-PrOH into appropriately
            labeled bottle and cap as before.

      D.    For each sample:

            1.     Shake can on paint shaker for 5 minutes.

            2.     Pipet 2 ml DMF into bottle.

            3.     Weigh to nearest 1 mg on balance.

            4.     Remove the bottle to hood and add about 0.4 to 0.6 grams of
                  sample dropwise using a transfer pipette.

                  a. Add 10 drops of sample and reweigh.

                  b.  Estimate how many additional drops of sample to add.

                  c.     In the hood add the sample dropwise until the required
                        amount is added.

                  d.     Reweigh to verify the amount added.

                  e.     Make note of the total number of drops of sample added
                        and the weight increase.

            5.     Cap as before.

            6.     Shake vigorously.

III. Screen Samples

      A.    Use the DMF as the  syringe rinse between every injection.

      B.    Flush a 10 /*! syringe with the pure DMF from its labelled bottle five times.
            Dispose of the waste into a 20 ml scintillation vial set aside for waste
            DMF.

      C.    Insert needle into DMF bottle all the way to the liquid level and repeatedly
            fill and empty the syringe to expel air.  Raise the needle stopping in the


Page 22

-------
      air space above the liquid.  Pull the plunger to the 1 /*! mark drawing in
      air.

D.    Lower the needle into the liquid to be injected and pull the plunger to the
      2 u\ mark.  Raise  the needle out of the liquid and advance the plunger to
      the 4  pi mark. Verify the loading of exactly  1 /J of liquid into the syringe
      before removal from the bottle.

E.    When it has been verified that the volume is accurate, remove the needle
      and syringe from  the bottle.

F.    Using the GC conditions outlined in Section IX of this  procedure, inject
      the contents of the syringe into the GC.

G.    Hit the start buttons on the GC and the integrator.

H.    Rinse the syringe with DMF at least five times disposing of the waste as
      before.

I.     Repeat steps A through H for the  DMF  + i-PrOH  + n-PrOH solution and
      the sample solutions.

J.    The following information should be noted for each sample:

      1.     The  presence of solids which made filling the syringe difficult and
            what .steps had to be used  to fill the syringe. Options include the
            following:

            a.     Allowing the solids to settle, insert the needle into the air
                   space.  Advance the plunger to the  1 //I  mark drawing in air
                   and  then carefully invert the sample and load the sample.

            b.     Centrifugation using a Dynac tabletop centrifuge  (or
                   comparable equipment)  at 70% maximum speed for five
                   minutes. The samples are  in long narrow tubes with screw
                   caps. After centrifugation, the  supernatant is transferred to
                   a small bottle and then capped.

            c.     Filtration, if used,  is a last resort. Cleaning the filter in any
                   quantitatively accurate way is difficult.

      2.     Approximate water content. For samples with little or no water,
            0.6 g of material can be used later when preparing for  analysis.
                                                                  Page 23

-------
                  Those with enough water to overwhelm the detector should be
                  prepared using 0.3 g, 0.2 g or 0.1 g as necessary.

            3.    Look for potential interference with i-PrOH elution. If the sample
                  contains i-PrOH or peaks that would interfere with its
                  quantification, select  an alternative internal  standard.  Examine the
                  elution peak of n-PrOH. If its elution position is free of
                  interference, use it instead.  In extreme cases ethanol may be
                  used.

IV.    Prepare Internal Standard

      A.    Make the internal standard  selection based on the screening results.

      B.    Since in most cases either i-PrOH or n-PrOH will be satisfactory, using
            the labelled syringe withdraw the approximate amount of internal
            standard needed (1 ml + 0.75 ml per sample).

V.    Prepare Blanks

      A.    Label one 10 ml serum bottle B-1  and a second one B-2.

      B.    Add 2.0 ml DMF with the appropriate pipettor to B-1.
                                                  •4
      C.    Weigh the bottle, using gloves or tissue when handling to prevent finger
            prints. Record the weight to the nearest 0.1  mg.

      D.    Using the Eppendorf pipettor add 200 n\ + 5 drops of the selected
            internal standard.

      E.    Weigh and record as before.

      F.    Seal well and shake  to mix.

      G.    Repeat steps B through F using B-2.

VI.    Prepare References

      A.    Label one 10 ml serum bottle R-1  and a second one R-2.

      B.    Follow steps B through E of the blank preparation procedure described
            above using the R-1  bottle.

      C.    With the  Eppendorf pipette  and a clean tip add 200 ^l deionized H2O.

Page 24

-------
      D.    Weigh, record the weight, seal well and shake.

      E.    Repeat steps B through D using R-2.

VII.   Prepare Samples

      A.    Label three 10 ml serum bottles for each sample with the unique sample
            identification number and add a different letter (A, B, or C) to each to
            make each bottle label unique.

      B.    Mix paint or coating for five minutes on the paint shaker if it was not
            mixed recently.

      C.    Follow steps B through E of the blank preparation procedure using the
            _A bottle.

      D.    Modify the total weight of coating to be added using the data noted
            during the screening.  The relationship of the number of drops to the
            weight noted during the sample screen preparation will help here.  Add
            the sample now.

      E.    Weigh, record the weight and cap the bottle immediately.

      F.    Mix the sample vigorously.

      G.    Repeat steps C through for the	B and	C bottles.

      H.    Repeat steps B through G for the other samples.

VIII.   Analyze Samples

      A.    Use B-1 as the syringe rinse between every injection.

      B.    Fill a 10 //I syringe with B-1 five times, disposing of the  liquid in the waste
            DMF vial.

      C.    Draw in 1 n\ of B-1 as described earlier.

      D.    Using the GC conditions outlined in section IX of this procedure, inject
            the contents of the syringe into the GC.

      E.    Hit the start buttons on both the GC and the integrator.
                                                                        Page 25

-------
      F.     Rinse the syringe with B-1 five times, regardless of what was injected to
            clear the syringe of potential solid deposits.

      G.     When the GC is ready for the next injection,  rinse the syringe again with
            B-1 five times.

      H.     Repeat steps C through G for each of the following: B-2, R-1, R-2,  and
            the three bottles (A, B, and C) prepared for the paints or coatings being
            analyzed.

IX.  '  GC Conditions

      A.     Use a Hewlett Packard 5890 A gas chromatograph (or comparable
            equipment) with a thermal conductivity detector (TCD).

      B.     Column.

            1.     Packing                                   Porapak Q

            2.     Dimensions                                10 ft-4 mm ID

            3.     Column Material     .                       Glass-1/4" OD

      C.     Use a Hewlett Packard 3357 Laboratory Automation System backed up
            by a  Hewlett Packard 3392A integrator or equivalent equipment.

      D.     GC parameters.

            1.     Flows

                  a.     Column                             45 cc/min

                  b.     Total (column + ret):                 112 cc/min

            2.     Oven

                  a.     Equilibrium time                      0.25 min

                  b.     Initial oven temperature                90°C
                        (for  a new column)                   70°C

                  c.     Initial time                           0.00 min

                  d.     Rate                                20°C/min

Page 26

-------
            e.     Rnal oven temperature                210°C



            f.     Final time                            24.00 min



      3.     Run length                                 30.00 min



      4.     Signal 1



            a.     Range                              4



            b.     Zero                                0.0



            c.     Attenuation                          0



      5.     Injector



            a.     Injector A temperature                220°C



      6.     Detector



            a.     Detector A TCD                      ON (+)



            b.     Detector temperature                 240°C



      7.     Purge ON



E.     Integrator parameters.



      1.     Zero                                      20



      2.     Attenuation  2                              4



      3.     Chart speed                               0.5 cm/min



      4.     Peak width                                 0.16



      5.     Threshold                                 5



      6.     Area reject                                 100



      7.     Time table



            a.     INTG # 10                           13.50 min
                                                                Page 27

-------
                  b.
                  INTG # 10
18.80 min
X.
      F.


      G.

      H.
      To start GC turn on oven temperature and detector.  When both lights
      are off, temperatures are at set points and the GC is ready.

      When finished with analyses, turn off the detector and the oven.

      GC maintenance.

      1.     Change the septum each day.  Remember to turn off the detector
            and the oven while making the change.

      2.     Make sure that the nut holding the septum in place is tightly
            turned.

      3.     Make sure the TCD is turned off for the night.
Data
      The following data should be recorded (I.S. is internal standard):
Sample
Weights
B1
B2
R1
R2
A
B
C
WtVial + DMF
(9)







+ I.S.
(g)







+ Water
(g)







+ Sample
(g)







Page 28

-------
      Where:
              Wt Vial + DMF
                       + IS

                    + Water

                  +  Sample
gross weight of vial and DMF solvent.
(Wt. Vial + DMF) + weight of internal
standard added.
(+ IS) + weight of water added to reference
sample R1 and R2.
(+ IS) + weight of paint or coating added to
each of A, B, or C for analysis
Sample Areas:
B1
B2
R1
R2
A
B
_C
Water Area







I.S. Area







DMF Area







      Where:
            Water Area  = G.C. area counts for water
            I.S. Area     = G.C. area counts for the internal standard
            DMF Area   = G.C. area counts for the DMF solvent
XI.    Calculations

      A,, = Respective areas

      W,,  = Respective weights
            VMFJ
                                                                      Page 29

-------
            Where:

            i = Chemical compound
            j = Blank or reference standard number (1 or 2)
            B = area of the water peak in the blank
            R = Response factor of water

      Calculations are best performed by computer on a spreadsheet in conjunction
      with an automated data acquisition system for analytical input from the gas
      chromatograph system.
XII.   Notes
      A.    Whenever a coating sample contains more than 40% H2O, a new sample
            should be prepared using 0.4 g, 0.2 g or 0.1 g of coating.

      B.    If some compound elutes late and interferes with subsequent runs,
            increase the Final Time of the chromatograph run sufficiently to elute the
            substance prior to the start of the next run.  A Final Time in the range of
            25 minutes is usually sufficient to purge the column.
Page 30

-------
DETERMINATION OF WATER CONTENT IN COATINGS BY KARL FISCHER TITRATION
(Adapted from ASTM D 4017-81)
METHOD:   The water content of coating materials is determined by dissolving the
            sample in a suitable solvent and titrating it directly with standardized Karl
            Fischer reagent to an electrometric end point.

EQUIPMENT:

I.     Glassware/Labware

      A.    Reagent bottle top volumetric dispensers for chemicals which must be
            maintained under anhydrous conditions.

II.     Syringes

      A.    A 100 n\ syringe with needle - one.

      B.    A 10 ml disposable syringe without needle, but equipped with a cap -
            one.

      C.    A  1 ml disposable syringe without needle, but equipped with a cap -
            one.

III.    Chemicals

      A.    Water, Type  II reagent grade conforming to ASTM Specification D1193 is
            best. (This calls for the use of distilled water with a conductivity of less
            than 1.0 //mho/cm at 25 °C.) Freshly deionized (Dl) water may also be
            used.

      B.    Karl Fischer Reagent (KFR), A solution which is available from suppliers
            of laboratory chemicals and which contains iodine, sulfur dioxide, and
            pyridine dissolved in ethylene glycol monomethyl ether.  Methanol is not
            generally  used because 1) it may  not dissolve many common resins, 2)
            methanol reacts with some resins to produce water, and 3) methanol
            may participate in water-forming esterification reactions in the presence
            of certain mineral acids (for example, in the formation of acetals and
            ketals from reaction  with  aldehydes and ketones), respectively, or may
            participate in side reactions with certain amines and siloxanes.

      C.    Pyridine, reagent grade.  Available from suppliers of laboratory
            chemicals.

                                                                       Page 31

-------
      D.    1-Ethyipiperidine, reagent grade.  Available from suppliers of laboratory
            chemicals.

IV.    Miscellaneous

      A.    Safety Equipment.  Auxiliary laboratory equipment should include
            provisions for hood ventilation of benchtop work space as well as
            ventilation of the general laboratory area, storage for chemical reagents,
            and storage for solvents and coatings and inks which present a
            flammability risk because of their volatile solvent content.

V.    Analytical Balance, Range: 0-20Q grams, Reproducibility: ±  0.1 mg.

VI.    Karl Fischer Apparatus, manual or automatic, encompassed by the detailed
      description given in ASTM Standard E 203-75 Test Method  for Water Using Karl
      Fischer Reagent.  Apparatus should be equipped with a 20-25 ml buret with 0.1
      ml divisions.  A 1  ml microburet with 0.01 divisions may be  obtained for analysis
      of samples containing less than 0.5%  water by weight.

PROCEDURE:

I.     General Preparation

      A.    Assemble equipment needed.

            1.      Check Karl Fischer Apparatus being used. Clean titration vessel if
                   needed with fresh pyridine and discard waste  liquid into a
                   dedicated and sealed waste disposal container (e.g., a 5 gallon
                   can with safety lid) which is stored  in the hood.  Perform
                   preliminary equipment checkout steps (following equipment
                   manufacturer's instructions).

            2.      Clean up work area, removing waste paper, solvent, chemical
                   reagent, or used labware materials.

            3.      Follow manufacturer's maintenance and setup instructions for the
                   Karl Fischer Apparatus being used.

      B.    Assemble Reagents.

            1.      Safety  Precautions. The Karl Fischer Reagent (KFR) used in  this
                   procedure contains four toxic compounds, namely iodine, sulfur
                   dioxide, pyridine, and ethylene glycol monomethyl ether.  1-
                   ethylpiperidine used as a catalyst for  the reagent should be

Page 32

-------
                  considered equally toxic to pyridine and therefore handled with the
                  same care.  All operations including dispensing of reagents should
                  be carried out in a hood.  Rubber gloves and a face shield should
                  be worn when handling and transferring chemicals.  Care must
                  exercised to avoid inhalation or skin contact.

            2.    MSDS (Material Safety Data Sheet) instructions.  All reagents used
                  should be supplied with MSDS forms which contain physical,
                  chemical, and safety guideline/emergency information. The MSDS
                  sheets should be available in the same laboratory area where the
                  equipment is used. Particular attention should be paid to the
                  recommended emergency procedures to be followed in the event
                  of spillage or accidental inhalation or physical contact with the
                  chemicals being used.

            3.    Maintain anhydrous conditions.  Follow instructions regarding
                  maintenance of purity and stability of reagent chemicals that are
                  provided by laboratory chemical suppliers.  Do not leave
                  containers open or exposed to excessive sunlight or other extreme
                  conditions of temperature, humidity, or in potential contact with
                  other reactive materials.  Use volumetric safety dispensers for
                  reagent bottles which are available from laboratory equipment
                  suppliers.

II.     Standardize Karl Fischer Reagent

      A.    Charge Karl Fischer Apparatus with reagents.

            1.    Charge a clean,  dry titration vessel (containing a magnetic stirring
                  bar) with pyridine to a level sufficient to cover the electrode sensor
                  tip  plus 1  ml of 1 -ethylpiperidine catalyst per 20 ml of pyridine.
                  The 5% volumetric proportion of the catalyst yields the  best
                  titration performance. This step may be accomplished
                  automatically by different apparatus but the net result should be
                  the same as outlined above.  The magnetic stirrer should be on at
                  this time to provide mixing of the titration vessel solution.  The
                  stirring rate should be vigorous  but the end point electrodes
                  should remain fully submerged in the solution and excess bubble
                  formation should be avoided.

            2.    Charge the Karl Fischer Reagent (KFR) dispensing buret with KFR
                  to a point where an initial reading can be taken of the volumetric
                  level. Follow manufacturer's instructions if automatic or digital
                  dispensing provisions are included in the apparatus being used.


                                                                         Page 33

-------
      B.     Pretitrate pyridine/1-ethylpiperidine solution to the end point indicated by
            the equipment manufacturer by adding KFR from the dispensing buret to
            cause the end point to hold for at least 30 seconds.  Record the
            volumetric level in the buret.

      C.     Adjust the Karl Fischer apparatus for maximum electrode sensitivity and
            minimum titration rate. Follow manufacturer's guidelines.

      D.     Weigh the water standard specimen.  Fill the 100 /d syringe with Type II
            reagent grade distilled water and weigh the syringe and water to the
            nearest 0.1 mg.  Record the weight in grams to the fourth decimal place.

      E.     Inject the water standard from the 100 ^l syringe into the titration vessel
            using the appropriate sample port in the Karl Fischer apparatus.  If
            necessary,  immediately replace the  sample port stopper to reseal the
            titration vessel.

      F.     Titrate the water standard to the end point using the KFR dispensing
            buret to cause the end point to hold for at least 30 seconds and record
            the KFR volumetric level.

      G.     Weigh the empty syringe to the nearest 0.1 mg  and record the weight in
            grams to the fourth decimal place.

      H.     Calculate F (=Karl Fischer titre) in grams H2O per ml of Karl Fischer
            Reagent used.

      I.     Calculation formula:

            F = J/P

            Where:

            F = KFR titre
            J = water added, grams
            P =  KFR used, ml

            The value of F should be calculated and recorded to four significant
            figures. Typical values of F are in the range of 0.004000 to 0.006000
            grams/ml corresponding to a  KFR titre volume, P of 12.5 to 8.33 ml,
            respectively for a 50 n\ (approximately 0.05 gram) water standard
            specimen or sample, J.
Page 34

-------
G.    Repeat the standardization until replicate values of F agree within 1%.
      Determine the mean of at least two such determinations.  Carry out
      calculations retaining at least one extra decimal figure beyond that of the
      acquired data.  Round off figures after final calculations.

Perform Screening Specimen Analysis

A.    Prepare the titration vessel and KFR buret as described in Steps II. A and
      II. B (except using the 25 ml buret for dispensing the KFR). Best results
      are obtained wrth fresh solvent solution, that is containing no previously
      titrated specimen in the vessel.

3.    Mix the coating or mk sample container thoroughly (e.g., in a bench
      scale paint shaker/mixer for a period of five mmutes_)

C.    Draw a sample of the coating  or ink using a 1  ml or 10 ml syringe (with
      cap) applying the specimen size guidelines given in Table  1:
Expected Water,
(%)
0.5 - 1.0
1 -2
3 - 10
10 -30
30 - 70
> 70
Approximate Specimen
Weight
(g)
5
2 - 5
1 - 2
0.4 - 1.0
0.1 -0.4
0.1
Approximate Titrant
Volume at 5 mg/ml titr
(ml)
5 - 10
10-20
10 -20
20-25
15-25
20
      Remove the syringe from the sample container, pull the plunger out a
      little further, wipe the excess material off the syringe, and replace the cap
      on the syringe tip.  Weigh the filled syringe  to the nearest 0.1  mg.
      Reseal the sample container.

      Inject the sample into the titration vessel containing the pretitrated
      pyndine solution (after removing the syringe cap) and immediately reseal
      the vessel, if necessary.  A high stirring rate should  be employed and at
      least 2 minutes (preferably 5 minutes) should be allowed for the sample
      to dissolve before starting the titration.  Pull the plunger back  out of the
      syringe and replace the cap.
                                                                   Page 35

-------
      F.    Reweigh the emptied syringe to the nearest 0.1 mg and calculate the
            specimen weight by difference.

      G.    Calculation:

            % water = (milliliters KFR used x F x 100)/(grams of sample)

IV.    Perform the Sample Analysis

      A.    Adjust the specimen size based on findings of the screening analysis and
            repeat the determination. The % water is determined by obtaining a
            duplicate set of measurements and taking the average of the two results,
            at a minimum.

      B.    If the calculation shows that 0.1  to 0.5 % water is present in the sample,
            repeat the procedure using a 1 gram specimen and a 1 ml microburet for
            KFR titration.

V.    Notes

      A.    Accuracy-Repeatability/Reproducibility

            1.    Repeatability - Two results, each the mean of duplicate
                  determinations, obtained  by the same operator on different days
                  should be considered suspect if they differ by more than 4.7 %
                  relative.

            2.    Reproducibility - Two results, each the mean of duplicate
                  determinations, obtained  by operators in different laboratories
                  should be considered suspect if they differ by more than 15.0 %
                  relative.

      B.    Interferences

            1.    The possibility of interfering side reactions or the formation of
                  byproducts which prevent accurate determination of water content
                  is always possible.  This method has been developed to minimize
                  those potential problems but if they are suspected,  it should be
                  reported.

            2.    It is essential to use care and follow precise procedures when
                  dissolving and mixing the specimen samples to obtain a
                  homogeneous and therefore representative sample of the coating
                  or ink.


Page 36

-------
      3.    All reagents used should be less than one year old and from a
            reliable laboratory reagent supplier.  Contamination of reagents
            and solutions may cause significant reproducibility problems.

C.    Maintenance

      1.    Contamination of the KFR burets, the reagent reservoir glassware,
            the titration vessel, and the coating of the sensing electrodes are
            all potential problems to inspect for and to prevent by performing
            periodic cleaning.

      2.    Clean the titration vessel by rinsing with fresh pyridine. Do not
            use methanol or other solvents.

      3.    Maintain anhydrous conditions (dryness) in the titration vessel by
            checking that drying tubes are in good condition and are tightly
            connected.  Replace desiccant when indicator color changes
            through half the tube.

      4.    Follow manufacturer's guidelines regarding electrode performance
            response and cleaning.

D.    Disposal of Waste

      1.    A separate waste container, such as a 5 gallon solvent can with lid
            should be maintained for spent or used reagents resulting from
            the use and cleaning of the titration vessel. This should  be
            emptied for periodic disposal as a hazardous laboratory  waste by
            a reputable disposal  organization.
                                                                  Page 37

-------
Page 38

-------
DETERMINATION OF PERCENT VOLATILE CONTENT IN COATING SAMPLES (Adapted
from ASTM D2369-81)
METHOD:   The percent volatile content (including water) as a weight fraction is
            determined by measuring the weight loss of a known sample quantity,
            which has been heated for one hour (60 minutes) at 110°C in a forced-
            draft oven.

EQUIPMENT:

I.     Glassware

      A.     Bottles of 50 ml capacity with a cap having an inert liner - two.

II.     Syringes and Pipets

      A.     5 ml disposable syringes without needle - one per sample.

      B.     A 10 ml pipette - one per sample.

III.    Chemicals

      A.     Deionized water (H20).

      B.     Toluene.

IV.    Miscellaneous

      A.     Aluminum foil weighing dishes - three per sample.

V.     Analytical  balance - Range: 0-200 grams, Reproducibility: ±0.1 mg

VI.    Vacuum Oven and Vacuum Pump

VII.    Desiccator

PROCEDURE:

I.     General Preparation

      A.     Assemble equipment needed.

      B.     Put toluene in one of the 50 ml bottles and deionized water in the other.


                                                                    Page 39

-------
      Preparation of Samples

      A.     Inscribe an identification mark on each of three aluminum weighing
            dishes (e.g. A, B, C) for each coating.  Remember to add an identifying
            mark to separate the coatings if more than one is under study.

      B.     Preheat the pans in the vacuum oven for 30 minutes at 110°C.

      C.     Place the dishes in a desiccator to cool to ambient temperature (about
            15 minutes).

      D.     Weigh the dishes and record the weight.

      E.     For each coating:

            1.     Shake can on  paint shaker for 5 minutes.

            2.     Add 3 ml  ±  1 ml of suitable solvent (toluene or water) to each of
                  the three weighing dishes using the 10 ml pipettor.

            3.     Using the information obtained from the screening of samples
                  prior to determination of water content, estimate volatile content
                  and proceed.

            4.     Add to each weighing dish 0.6 ml of coating if it has a volatile
                  content less than 50 weight % or 0.8 ml if it has a volatile content
                  greater than 50 weight %.

                  a.    Use a 5 ml disposable  syringe.

                  b.    Fill the syringe with coating.

                  c.    Weigh the filled syringe to the nearest 0.1 mg and record
                        the weight.

                  d.    Add an aliquot of the sample to the first dish, reweigh the
                        syringe and record the weight.

                  e.    Add an aliquot of the sample to the second dish, reweigh
                        the syringe and record the weight.

                  f.     Repeat this for the third dish.
Page 40

-------
            5.     Heat the prepared weighing dishes in the vacuum oven for 60
                  minutes at 110 ± 5°C with the vacuum valve cracked open but the
                  vent wide open (indicated vacuum barely detectable on vacuum
                  gauge, corresponding to 0.1 to 1" Hg vacuum).

            6.     Remove the dishes from the oven and place them immediately in a
                  desiccator,  cool to ambient temperature  about 15 minutes and
                  weigh them to the nearest 0.1  mg.
111.
Data
      The following data is recorded:
Samples
A
B
C
Pan Wt.
(9)



Initial Wt.
Sample +
Syr
(9)



Final Wt.
Sample +
Syr.
(g)



Final Wt.
Pan +
Sample
(g)



Final Wt.
Sample.
(g)



VI.
Calculations

%VOC = Percent volatile matter in coating.

W, = Weight of the ith item.
              -W
              ~ '*
                 InltSmpb+Syr
                                         100
                         'Sa/npfe
                                                                      Page 41

-------
DETERMINATION OF DENSITY OF PAINT, VARNISH, LACQUER, AND RELATED
PRODUCTS (Adapted from ASTM D1475-60 (Reapproved 1980))
METHOD:   The density of the liquid at 25°C expressed in gm/ml and Ibs/gal is
            determined by the following procedure.  The accurately known absolute
            density of distilled (or deionized) water is used to calibrate the volume of
            a container.  The same container is filled with sample liquid and weighed.
            All measurements are taken after the container and contents equilibrate
            at 25°C. .

EQUIPMENT:

I.     Glassware

      A.    Container for boiling and storing distilled water.

      B.    Pycnometer - three per sample.

II.  Syringes and Pipettes

      A.    A 25 ml pipette - three per sample.

III.    Chemicals

      A.    Deionized water (H2O).
                                     •

IV.    Miscellaneous

      A.    Constant temperature bath held at 25°C ±0.1°C.

      B.    Desiccator.

V.     Analytical Balance - Range: 0-200 grams, Reproducibility: ±0.1 mg

PROCEDURE:

I.     General  Preparation

      A.    Assemble equipment needed.

      B.    Freshly boil deionized water for use in the calibration step.
                                                                      Page 43

-------
II.     Determine Tare Weight of the Dry Pycnometers

      A.    Clean and dry the pycnometers and bring each of them to a constant
            weight.

            1.     Strong acid or strong base cleaners and nonresidual solvents may
                  be used with glass containers.

            2.     For maximum accuracy, continue rinsing, drying, and weighing
                  until the difference between two successive weighing does not
                  exceed 0.001% of the weight of the container (0.0003 g for a 30 g
                  pycnometer).

            3.     Fingerprints  on the container will change the weight and must be
                  avoided.

            4.     Record the weight. This and all subsequent references to
                  recording of information refer to the written account kept in the
                  data books for environmental samples.

III.    Calibration of Pycnometers

      A.    Fill the pycnometers with the freshly boiled distilled water  (or freshly
            boiled deionized water) at a temperature below 25°C.

      B.    Immediately remove excess overflowed water or water held in
            depressions by wiping dry with absorbent material.

      C.    Avoid occluding air bubbles in the container.

      D.    Bring the container and contents to 25°C using the constant temperature
            bath.

      E.    Remove the excess overflow by wiping carefully with absorbent material.
            Avoid wicking of water out of the orifice.

      F.    Dry the outside of the pycnometers.

      G.    Immediately weigh the filled pycnometers to the nearest
            0.001% of its weight (0.0003 g for a 30 g pycnometer).

      H.    Record the weight.
Page 44

-------
IV.
V.
Density of the Sample

A.    For each coating:

      1.     Shake can on paint shaker for 5 minutes if that has not been done
            recently.

B.    Repeat III using the sample instead of distilled water and using the 25 ml
      pipette to load the pycnometer.

Data

The following data is  recorded:
Samples
A
B
C
Pyc#



Pyc Wt.
(g)



Pyc +
Sample Wt.
(g)



Pyc Vol.
(ml)



VI.    Calculations.
      For the calibration:
        y.
          0.997072

      Obtain the mean of at least three determinations.

      Where:

      V = volume of the pycnometer, ml.

      N = weight of the pycnometer and water, g.

      M = weight of the dry pycnometer, g.

      0.997072 = absolute density of water at 25°C, g/ml
                                                                       Page 45

-------
      For the density determination:
                   and  D=Dm*K
     Where:



     Dm = density, g/ml and D = density, Ib/gal.



     W = weight of the filled pycnometer, g.



     w = weight of empty pycnometer, g.



     V = volume determined above, ml.



     K = 8.345404
Page 46

-------
DETERMINATION OF DICHLOROMETHANE AND 1,1,1-TRICHLOROETHANE IN PAINTS
AND COATINGS BY DIRECT INJECTION INTO A GAS CHROMATOGRAPH (Adapted
from ASTM D4457-85)
METHOD:   The percent dichloromethane and 1,1,1-trichloroethane content of
            coating material is determined by dissolving the sample in a suitable
            solvent and injecting it directly into a gas chromatograph. An internal
            standard is added to the sample in a known quantity before
            chromatography-for calibration.

EQUIPMENT:

I.     Glassware/Labware

      A.    Reagent dispensing buret, 100 ml capacity - one.

      B.    20 ml scintillation vials with teflon inverted cone cap seals - five per
            sample.

      C.    25 ml vials capable of  being septum sealed - six + four per sample.

            1.    Teflon coated septa - six + four per sample.

            2.    Aluminum caps - six + four per sample.

II. Syringes and pipettes

      A.    20 ml disposable syringes with needle - two.

      B.    10 ml disposable syringes with needle - six.

      C.    10 ml Eppendorf or equivalent volumetric pipettors with eight disposable
            tips - one.

      D.    Transfer pipettes for dropwise addition of samples.

      E.    A 10 n\  syringe with needle.

III.    Chemicals

      A.    Anhydrous N,N-Dimethylformamide (DMF), reagent grade.
                                                                     Page 47

-------
      B.     Anhydrous 1-propanol (or 1-PrOH, n-propanol), gas chromatography
            spectrophotometric grade.

      C.     Anhydrous 2-propanol (or 2-PrOH, isopropanol), gas chromatography
            spectrophotometric grade.

      D.     1,1,1 - Trichloroethane (or TCE, methyl chloroform), reagent grade (see
      E.     Dichloromethane (or DCE, methylene chloride), reagent grade   (see E).

      F.     Halogenated hydrocarbon stabilizers - Commercial grades of TCE and
            DCE contain stabilizers. Obtain the same solvent used in the coating or
            find the type and quantity of stabilizer used and add it to the pure
            solvent.
                                               o
      G.     Carrier gas, Nitrogen (NJ of 99.995% or higher purity. High purity
            Helium may also be  used.

IV.    Miscellaneous

      A.     Rubber tubing - 18 inches.

      B.     Bottle cap crimper.

      C.     Safety Equipment. Auxiliary laboratory equipment should include
            provisions for hood ventilation of benchtop work space as well as
            ventilation of the general laboratory area, storage for chemical reagents,
            and storage for solvents and coatings and inks which  present a
            flammability risk because of their volatile solvent content.

V.    Analytical Balance, Range:  0-200 grams, Reproducibility: ±0.1 mg.

VI.    Gas Chromatograph (GC) - e.g., HP 5890 A

      A.     Thermal conductivity detector.

      B.     Porapak column - 4  ft.

PROCEDURE:

I.     General Preparation

      A.     Assemble equipment needed.


Page 48

-------
      1.     Clean up work area, removing waste paper, solvent, chemical
            reagent, or used labware materials.

B.    Assemble reagents.

      1.     Safety Precautions. All operations including dispensing of
            reagents should be carried out in a hood.  Rubber gloves and a
            face shield should be worn when handling and transferring
            chemicals. Care must exercised to avoid inhalation or skin
            contact.

      2.     MSDS (Material Safety Data Sheet) instructions. All reagents used
            should be supplied with  MSDS forms which contain physical,
            chemical,  and safety guideline/ emergency information. The
            MSDS sheets should be available in the same laboratory area
            where the equipment is used.  Particular attention  should be paid
            to the recommended emergency procedures to be followed in the
            event of spillage or accidental inhalation or physical contact with
            the chemicals being used.

      3.     Maintain anhydrous conditions. Follow instructions regarding
            maintenance of purity and stability of reagent chemicals that are
            provided by laboratory chemical suppliers. Do not leave
            containers open or exposed to excessive sunlight  or other extreme
            conditions of temperature, humidity, or in potential contact with
            other reactive materials.  Use volumetric safety dispensers for
            reagent bottles which are available from laboratory equipment
            suppliers.

C.    Fill a 20 ml scintillation vial with dichloromethane, cap it and label the vial
      "DCM."  Fill a 20 ml scintillation vial with 1,1,1-trichloroethane, cap it and
      label it TCE."

D.    One 20  ml syringe with needle  and a 20 ml scintillation vial should be
      labeled "1-PrOH" and another 20 ml syringe with needle and 20 ml -
      scintillation vial should be labeled "2-PrOH".  Set aside one 20 ml
      scintillation vial, label it "Waste DMF," loosely cover it with a septum and
      cap, but do not seal it.

E.    If the bottles containing anhydrous DMF, anhydrous  1-PrOH and
      anhydrous 2-PrOH are not slightly pressurized relative to the surrounding
      atmosphere then that must be done now.  (This is needed to facilitate the
      withdrawal of material.)
                                                                  Page 49

-------
            1.     Attach rubber tubing to N2 supply manifold and close the other
                  outlets.

            2.     Turn on the N2 to a very gentle flow.

            3.     To pressurize the DMF bottle, wearing gloves remove the needle
                  from the DMF syringe and insert the needle base in the open end
                  of the tubing. (TAKE GREAT CARE to hold the needle in the tube
                  firmly or it will be released at this point like a projectile.)

            4.     Insert the needle in the septum of the DMF bottle and invert the
                  bottle.

            5.     When the bubbles decrease, right the bottle and remove  the
                  needle.
                                                                        6
            6.     Replace the needle on the DMF syringe.

            7.     Repeat the procedure for 1-PrOH and 2-PrOH using the appro-
                  priate needle if needed.

            8.     Turn off the N2 flow.

      F.     Check that the 100 ml dispensing buret is mounted on the DMF reagent
            bottle  and that sufficient DMF reagent is present to do the analysis
            (approximately 20 ml for each 25 ml vial used  in screening and analysis).

      Preparation of Screening Samples

      A.     Label the 25 ml vials capable of being septum sealed that are needed for
            this step. One should be labeled "DMF," another "DMF + 1-PrOH  + 2-
            PrOH." Additional vials,  one for each sample, should be labeled using
            the appropriate laboratory sample number.

      B.     Weigh the vial labeled "DMF" to 0.1  mg, dispense 17 ml DMF into the
            bottle  and weigh the vial again to determine the amount of DMF added.
            Place  septum-teflon side down-on top of the vial, cover both with metal
            cap and crimp.

            1.     The action of the  crimping tool should be gentle but firm.

            2.     The cap should not move if sealed properly.

            3.     If it is loose, rotate the vial turn and crimp  again.
Page 50

-------
      4.     Repeat until tightly sealed.

C.    Dispense 17 ml DMF, then pipet 2 ml + 10 drops 1-PrOH and 2 ml  + 10
      drops 2-PrOH into appropriately labeled vial obtaining weights of each of
      the three components added to 0.1 mg and cap as before.

D.    For each screening sample:

      1.     Shake can on paint shaker for 5 minutes.

      2.     Dispense 17 ml DMF into the sample vial.

      3.     Weigh to the nearest 1 mg on the analytical balance.

      4.     Remove the bottle to hood and add about 4 ml of sample
            dropwise using a 10 ml disposable syringe.

            a.     Add 10 drops of sample using a transfer pipette and
                  reweigh.

            b.     Estimate how many additional drops of sample to add  to
                  bring the total sample addition amount up to 5 g.

            c.     In the hood add the sample dropwise until the required
                  amount is added.

            d.     Reweigh to verify the amount added.

            e.     Make note of the total number of drops of sample added
                  and the weight  increase.

      5.     Cap as before.

      6.     Shake vigorously.

Screen samples

A.    Use the DMF from the 25 ml septum sealed vial as the syringe rinse
      between every injection.

B.    Flush a 10 n\ syringe with the pure  DMF from its  labelled bottle five times.
      Dispose of the waste into a 20 ml scintillation vial set aside for waste
      DMF and cover  loosely with a septum and cap.
                                                                Page 51

-------
      C.    Insert needle into DMF vial but stop in the air space above the liquid.
            Pull the plunger to the 1 /il mark drawing in air and then lower the needle
            into the liquid and pull the plunger to the 2 n\ mark.  Raise the needle out
            of the liquid and advance the plunger to the 4 //I mark.  Verify the loading
            of exactly 1 n\ of liquid into the syringe before removal from the bottle.

      D.    When it has been verified that the volume is accurate, remove the needle
            and syringe from the vial.

      E.    Using the GC conditions outlined in the GC section of this procedure,
            inject the contents of the syringe into the GC.

      F.    Hit the start buttons on the GC and the integrator.

      G.    Rinse the syringe with DMF at least five times disposing of the waste as
            before.

      H.    Repeat steps A through F for the DMF + 1-PrOH + 2-PrOH solution and
            the sample solutions.  If there is  any doubt of the elution times of 1-PrOH
            or 2-PrOH, the pure components can be injected into the gas
            chromatograph for positive determination.

      I.     The following information should be noted for each sample:

            1.    The presence of solids which made filling the syringe difficult and
                  what steps had to be used to fill the syringe, including:

                  a.     Allowing the solids  to settle, insert the needle into the air
                        space.  Advance the plunger to the 1 fj\ mark drawing in air
                        and then carefully invert the sample and load the sample.

                  b.     Centrifugation using a laboratory tabletop  centrifuge (or
                        comparable equipment) at 70% maximum speed for five
                        minutes. The samples are in long narrow tubes with screw
                        caps.  After centrifugation, the supernatant is transferred to
                        a small bottle and it is then capped.

                  c.     Filtration if used is a last resort. Cleaning the filter in any
                        quantitatively accurate way is difficult.

            2.    Approximate  exempt solvent content.  For samples with little or no
                  exempt solvent present, 5 g of material can be used later when
                  preparing for analysis.  Those with enough dichloromethane or
Page 52

-------
                   1,1,1-trichloroethane to overwhelm the detector should be
                   prepared using 3 g, 2 g, or 1 g as necessary.

             3.     Look for potential interference with 1-PrOH elution.  If the sample
                   contains 1 -PrOH or peaks that would interfere with its
                   quantification, select an alternative internal standard.  Examine the
                   eJution peak of 2-PrOH.  If its elution position is free of
                   interference, use it instead.  In extreme cases, ethanol, other
                   alcohols,  esters, or hydrocarbons may be used.

IV.    Prepare internal standard (IS)

      A.     Make the internal standard selection based on the screening results.

      B.     Since in most cases either 1 -PrOH or 2-PrOH will  be satisfactory, use the
             labeled syringe to withdraw the approximately 20 ml amount of internal
             standard needed (10 ml  + 7.5 ml per sample) and charge it to the
             labeled 20 ml scintillation vial.

V.    Prepare Blanks

      A.     Label one 25 ml vial B-1  and a second one B-2.

      B.     Add 17 ml DMF from the reagent dispenser to B-1.

      C.     Weigh the bottle, using gloves or tissue when handling to prevent finger
             prints.  Record the weight to five decimal places.

      D.     Using the Eppendorf (or equivalent) pipettor add 2.5 ml of the selected
             internal standard.

      E.     Weigh and record as before.

      F.     Seal well and shake to mix.

      G.     Repeat steps B through F using B-2.

VI.    Prepare References

      A.     Label one 25 ml vial R-1 and a second  one R-2.

      B.     Follow steps B through E of the  blank preparation procedure using the
             R-1 bottle.
                                                                         Page 53

-------
      C.    With the Eppendorf (or equivalent) pipette and a clean tip add 2.7 ml of
            dichloromethane.

      D.    Weigh and record the weight.

      E.    With the Eppendorf (or equivalent) pipette and a clean tip add 2.7 ml of
            1,1,1 -trichloroethane.

      F.    Weigh and record the weight.

      G.    Seal well and shake.

      H.    Repeat steps B through G using R-2.

VII.   Prepare Samples

      A.    Label three 25 ml vials for each sample with the unique sample
            identification number and add a different letter (A, B, or C) to each to
            make each bottle.

      B.    Mix paint or coating for five minutes on the paint shaker if it was not
            mixed recently.

      C.    Follow steps B through E of the blank preparation procedure using the
            _A bottle.

      D.    Modify the total weight of coating to be added using the data noted
            during the screening.  The relationship of the number of drops to the
            weight noted during the sample screen preparation  will help  here.  Add
            the sample now using a fresh 10 ml disposable syringe.

      E.    Weigh, record the weight and cap the bottle immediately.

      F.    Mix the sample vigorously.

      G.    Repeat steps C through F for the	B and	C bottles.

      H.    Repeat steps B through G for the other samples.

VIII.   Analyze Samples

      A.    Use B-1 as the syringe rinse between every injection.
Page 54

-------
      B.    Fill a 10 fj,\ syringe with B-1 five times disposing of the liquid in the waste
            DMF vial.

      C.    Draw in 1 /d of B-1 as described earlier.

      D.    Using the GC conditions outlined in the GC section of this procedure,
            inject the contents of the syringe into the GC.

      E.    Hit the start buttons on both the GC and the integrator.

      F.    Rinse the syringe with B-1 five times regardless of what was injected to
            clear the syringe of potential solid deposits.

      G.    When the GC is ready for the next injection, rinse the syringe again with
            B-1 five times.

      H.    Repeat steps C through G for each of the following:  B-2, R-1, R-2, and
            the three bottles (A, B, and C) prepared for the paints or coatings being
            analyzed.

IX.    GC Conditions

      A.    Use a Hewlett Packard 5890 A gas chromatograph (or comparable
            equipment) with a thermal conductivity detector (TCD).

      B.    Column.

            1.     Packing                                    Porapak R

            2.     Dimensions                                 4 ft-4 mm ID

            3.     Column Material                            Glass-1/4" CD

      C.    GC parameters.

            1.     Flows

                  a.    Column                              45 cc/min

                  b.    Total (column + ref.)                  100 cc/min

            2.     Oven

                  a.    Equilibrium time                       0.25 min


                                                                       Page 55

-------
                  b.     Initial oven temp.                     90°C
                        (for a new column)                   70°C

                  c.     Initial time                           0.00 min

                  d.     Rate  20°C/min

                  e.     Final oven temperature                210°C

                  f.     Final time                           24.00 min

            3.     Run length                                30.00 min

            4.     Signal 1

                  a.     Range                               4

                  b.     Zero                                30.0

                  c.     Attenuation                          0

            5.     Injector

                  a.     Injector A temperature                220°C

            6.     Detector

                  a.     Detector  A TCD                      ON (+)

                  b.     Detector  temperature                 240°C

            7.     Purge       ON

      D.    Integrator parameters.

            1.     Zero                                      20

            2.     Attenuation 2                               4

            3.     Chart speed                               0.5 cm/min

            4.     Peak width                                0.16

            5.     Threshold                                 5
Page 56

-------
X.
            6.     Area reject

            7.     Time table
                  a.
                  b.
                  INTO # 10
                  INTO # 10
                                                      100
13.50 min.
18.80 min.
      E.    To start GC turn on oven temperature and detector. When both lights
            are off, temperatures are at set points and the GC is ready.

      F.    When finished turn off the detector and the oven.

      G.    GC maintenance.

            1.     Change the septum each day.  Remember to turn off the detector
                  and the oven while making the change.

            2.     Make sure that the nut holding the septum in place is tightly
                  turned.
            3.     Make sure the TCD is turned off for the night.
Data.
      The following data should be recorded:
Sample
Weights
B1
B2
R1
R2
A
B
C
Wt
Vial + DMF
(9)







+ I.S.
(g)







+ dichloro-
methane
(g)







+ 1,1,1-
trichloro-
ethane
(g)







+ Sample
(g)







                                                                      Page 57

-------
      Where:
                    Wt Vial + DMF =
                             + IS =

                 + dichloromethane =

             + 1,1,1-trichloroethane =
                         + Sample =
gross weight of vial and DMF solvent.
(Wt Vial + DMF)  + weight of internal
standard added.
(4- IS) + weight of dichloromethan added
to the reference samples R1 and R2.
(+ dichloromethane)  + weight of 1,1,1-
trichloromethane added to reference
samples R1 and R2.
(+ IS) -i- weight of paint or coating sample
added to each of A, B, or C for analysis.
Sample
Areas
B1
B2
R1
R2
A
B
__C
Dichloro-
. methane
Area







1,1,1 -tri-
chlorometh-
ane Area







I.S. Area







DMF Area







      Where:
                    Dichloromethane Area =

              1,1,1 -Trichlorormethane Area =

                                 IS Area =

                               DMF Area =
      G.C. Area counts for
      dichlormethane.
      G.C. Area counts for 1,1,1-trichloro-
      methane.
      G.C. Area counts for the internal
      standard.
      G.C. Area counts for the DMF
      solvent.
XI.    Calculations
      A,, = Respective areas
Page 58

-------
W,, = Respective weights



  =  AESJ
  O    *•*•*
-------
      Calculations are best performed by computer on a spreadsheet in conjunction
with an automated data acquisition system for analytical input from the gas
chromatograph system.

Xll.   Notes

      A.    Whenever a coating sample contains more than 40% of either
            dichloromethane or 1,1,1-trichloroethane, a new sample should be
            prepared using 3 g, 2 g, or 1  g of coating.

      B.    If some compound elutes late and interferes with subsequent runs,
            increase the Rnal Time sufficiently to elute the substance prior to the
            start of the next run. A Final Time in the range of 25 minutes is usually
            sufficient to purge the column.

      C.    Accuracy-Repeatabiiity/Reproducibility for 1,1,1-trichloroethane.

            1.     Repeatability - Two results, each the mean of duplicate
                   determinations, obtained by the same operator on different days
                   should be considered suspect if they differ by more than 3.0%
                   relative.

            2.     Reproducibility - Two results,  each the mean of duplicate
                   determinations, obtained by operators in different laboratories
                   should be considered suspect if they differ by more than 8.1%
                   relative.

      D.    Accuracy-Repeatability/Reproducibility for Dichloromethane.

            1.     Repeatability - Two results, each the mean of duplicate
                   determinations, obtained by the same operator on different days
                   should be considered suspect if they differ by more than 3.0%
                   relative.

            2.     Reproducibility - Two results,  each the mean of duplicate
                   determinations, obtained by operators in different laboratories
                   should be considered suspect if they differ by more than  17.9%
                   relative.
Page 60

-------
E.    Interferences.

      1.     The possibility of interfering side reactions or the formation of
            byproducts which may prevent accurate determination of
            dichloromethane and 1,1,1-trichloroethane content is always
            possible.  This method has been developed to minimize those
            potential problems but  if they are suspected, it should  be reported.

      2.     It is essential to  use care and follow precise procedures when
            dissolving and mixing the specimen samples to obtain  a
            homogeneous and therefore representative sample of the coating
            or ink.

      3.     All reagents used should be less than one year old and from a
            reliable  laboratory reagent supplier.  Contamination of  reagents
            and solutions may cause significant r»eproducibility problems.

F.    Maintenance.

      1.     Contamination of the glassware/labware, syringes and pipettes,
            reagent bottles,  and dispensers are all potential problems to
            inspect  for and to prevent by performing periodic cleaning.

      2.     Maintain anhydrous conditions (dryness) in reagent vessels and
            glassware/labware by keeping seals and lids in place and tightly
            connected.

      3.     Follow manufacturer's guidelines regarding reagent bottle and
            syringe/pipettor/dispenser handling and cleaning.

G.    Disposal of waste.

      1.     A separate waste container,  such as a 5 gallon solvent can with lid
            should be maintained for spent or used reagents from  use and
            cleaning of the titration vessel. This should be emptied for
            periodic disposal as a hazardous laboratory waste by a reputable
            disposal organization.
                                                                  Page 61

-------
Page 62

-------
                                 SECTION 5

     REFERENCE METHOD 24A-DETERMINATION OF VOLATILE MATTER
    CONTENT AND DENSITY OF PRINTING INKS AND RELATED COATINGS
1.     Applicability and Principle

      1.1    Applicability.  This method applies to the determination of the volatile
            organic compound (VOC) content and density of solvent-borne (solvent
            reducible) printing inks or related coatings.

      1.2    Principle.  Separate procedures are used to determine the VOC weight
            fraction and density of the coating and the density of the solvent in the
            coating.  The VOC weight fraction is determined by measuring the weight
            loss of a known sample quantity which has been heated for a specified
            length of time at a specified temperature. The density of both the  coat-
            ing and solvent are measured by a standard procedure. From this
            information the  VOC volume fraction is calculated.

2.     Procedure

      2.1    Weight Fraction VOC.

            2.1.1  Apparatus.

                  2.1.1.1      Weighing Dishes. Aluminum foil, 58 mm in diameter
                             by 18 mm high, with a flat bottom. There must be at
                             least three weighing  dishes per sample.

                  2.1.1.2      Disposable Syringe.  5ml.

                  2.1.1.3      Analytical Balance. To measure to within 0.1 mg.

                  2.1.1.4      Oven.  Vacuum oven capable of maintaining a
                             temperature of 120 ±2°C and an absolute pressure
                             of 510 ±51 mm Hg for 4 hours. Alternatively,  a
                             forced draft oven capable of maintaining a tempera-
                             ture of 120 ±2°C for 24 hours.

            2.1.2  Analysis. Shake or mix the sample thoroughly to assure that all
                  the solids are completely suspended.  Label and weigh to the
                  nearest 0.1 mg  a weighing dish and record this weight (MX1).
                                                                     Page 63

-------
                  Remove a sample of the coating using a 5-ml syringe without a
                  needle.  Weigh the syringe and sample to the nearest 0.1 mg and
                  record this weight (MCY1).  Transfer 1 to 3 g of the sample to the
                  tared weighing dish.  Reweigh the syringe and sample to the
                  nearest 0.1 mg and record this weight (MCY2). Heat the weighing
                  dish and sample in a vacuum oven  at an absolute pressure of 510
                  ±51 mm Hg and a temperature of 120 ±2°C for 4 hours.
                  Alternatively, heat the weighing dish and sample in a forced draft
                  oven at a temperature of 120 ±2°C for 24 hours.  After the
                  weighing dish has cooled, reweigh it to the  nearest 0.1  mg and
                  record the weight (M^.  Repeat this procedure for a total of three
                  determinations for each sample.

      2.2   Coating Density. Determine the density of the ink or related coating
            according to the procedure outlined in ASTM D1475-60 (Reapproved
            1980), (incorporated by reference - see §60.17).

      2.3   Solvent  Density.  Determine the density of the solvent according to the
            procedure outlined in ASTM D1475-60 (Reapproved 1980). Make a total
            of three determinations for each coating. Report the density  D0 as the
            arithmetic average of the three determinations.
3.     Calculations
      3.1    Weight Fraction VOC. Calculate the weight fraction volatile organic
            content W0 using the following equation:
            Report the weight fraction VOC W0 as the arithmetic average of the three
            determinations.

      3.2   Volume Fraction VOC.  Calculate the volume fraction volatile organic
            content V0 using the following equation:
                                       W  D
                                   y =_ll£Jl£                         Eq. 24A-2
Page 64

-------
4.     Bibliography

      4.1    Standard Test Method for Density of Paint, Varnish, Lacquer, and Related
            Products.  ASTM Designation D1475-60 (Reapproved 1980).

      4.2    Teleconversation.  Wright, Chuck, Inmont Corporation with Reich, R. A.,
            Radian Corporation.  September 25, 1979.  Gravure Ink Analysis.

      4.3    Teleconversation.  Oppenheimer, Robert, Gravure Research Institute with
            Burt, Rick, Radian Corporation, November 5, 1979. Gravure  Ink Analysis.
                                                                       Page 65

-------
Page 66

-------
                                 SECTION 6

     STANDARD OPERATING PROCEDURES FOR METHOD 24A ANALYSES


APPLICATION OF EPA METHOD 24A

      EPA Method 24A is performed on publication rotogravure inks for New Source
Performance Standard (NSPS) compliance.  These inks are viscous, solvent-based
formulations which frequently must be transferred from sample containers with small or
large blade spatulas. Although not needed to measure weight fraction VOC, to
determine the volume fraction VOC, it is necessary to have a separate sample of the
ink solvent for density measurement.  There are no dispersing agent which are
presently recommended to prevent incorporation of solvent compounds into the
hardened dry residue remaining after oven treatment performed for VOC weight
percent measurement.

VOC DETERMINATION

      The determination of the VOC weight fraction is otherwise clearly described in
the method and SOP for Method 24 given in Section  4. No additional instructions are
considered necessary.

DENSITY DETERMINATION

      In the  process of making ink density determinations by ASTM D1475-60
(Reapproved 1980), it is critical to carefully pack the heavy coating into the
pycnometer using small and large blade spatulas without allowing air bubbles or void
spaces to be formed. It is also important to thoroughly clean the outside of the
pycnometer and carefully avoid weighing errors which can significantly affect the
accuracy of the VOC weight fraction determination.  Otherwise, SOP guidance for use
of ASTM D1475-60 with  Method 24, given in Section 4, is applicable for use with
Method 24A as well.
                                                                     Page 67

-------
Page 68

-------
                                 SECTION 7

                               BIBLIOGRAPHY

ASTM Procedures in Order of ASTM Number
 1.    Standard Specification for Industrial Grade Toluene, ASTM D362-S4. 1990
      Annual Book of ASTM Standards, Volume 6.03. Philadelphia, PA, 1990.

 2.    Standard Specification for Reagent Water, ASTM D1193-77.  1990 Annual Book
      of ASTM Standards, Volume 6.03.  Philadelphia, PA, 1990.

 3.    Standard Test Method for Water in Volatile Solvents (Fischer Reagent Titration
      Method), ASTM D1364-87. 1990 Annual Book of ASTM Standards, Volume
      6.03.  Philadelphia, PA, 1990.

 4.    Standard Test Method for Density of Paint, Varnish, Lacquer, and Related
      Products, ASTM D1475-85.  1990 Annual Book of ASTM Standards, Volume
      6.01.  Philadelphia, PA, 1990.

 5.    Standard Test Method for Volatile Content of Coatings, ASTM D2369-87. 1990
      Annual Book of ASTM Standards, Volume 6.01. Philadelphia, PA, 1990.

 6.    Standard Test Method for Volume Nonvolatile Matter in Clear or Pigmented
      Coatings, ASTM D2697-86.  1990 Annual Book of ASTM Standards, Volume
      6.01.  Philadelphia, PA, 1990.

 7.    Standard Guide for Determining Volatile and Nonvolatile Content of Paint and
      Related Coatings, ASTM D2832-83.  1990 Annual Book of ASTM Standards,
      Volume 6.01.  Philadelphia, PA, 1990.

 8.    Standard Specification for 2-Ethoxyethyl Acetate (99%  Grade), ASTM D3728-88.
      1990 Annual Book of ASTM Standards, Volume 6.03.  Philadelphia, PA, 1990.

 9.    Standard Test Method for Water Content of Water-Reducible Paints by Direct
      Injection Into a Gas Chromatograph, ASTM D3792-86.  1990 Annual Book of
      ASTM Standards, Volume 6.01.  Philadelphia, PA,  1990.

10.    Standard Practice for Sampling Liquid Paints and Related Pigmented Coatings,
      ASTM D3925-81.  1990 Annual Book of ASTM  Standards, Volume 6.01.
      Philadelphia, PA, 1990.
                                                                     Page 69

-------
11.    Standard Practice for Determining Volatile Organic Compound (VOC) Content of
      Paints and Related Coatings, ASTM D3960-89.  1990 Annual Book of ASTM
      Standards, Volume 6.01.  Philadelphia, PA, 1990.

12.    Standard Practice for Interlaboratory Testing of Paint and Related Materials,
      ASTM D3980-88.  1990 Annual Book of ASTM Standards, Volume 6.01.
      Philadelphia, PA, 1990.

13.    Standard Test Method for Water in Paints and Paint Materials by Karl Fischer
      Method, ASTM D4017-88.  1990 Annual Book of ASTM Standards, Volume
      6.01.  Philadelphia, PA, 1990.

14.    Manual Sampling of Petroleum and Petroleum Products, ASTM D4057-81.  1990
      Annual Book of ASTM Standards, Volume 5.03.  Philadelphia, PA, 1990.

15.    Standard Test Method for Determination of Dichloromethane and 1,1,1-
      Trichloroethane in Paints and Coatings by Direct Injection into a Gas
      Chromatograph, ASTM D4457-85.  1990 Annual Book of ASTM Standards,
      Volume 6.03.  Philadelphia, PA, 1990.

16.    Specification for Gravity Convection and Forced-Ventilation Ovens, ASTM E145-
      68.  1990 Annual Book of ASTM Standards, Volume 14.02. Philadelphia, PA,
      1990.

17.    Practice for Determining the Precision Data of ASTM Methods for Analysis and
      Testing of Industrial Chemicals, ASTM E180-85.  1990 Annual Book of ASTM
      Standards, Volume 15.05.  Philadelphia,  PA, 1990.

18.    Standard Test Method for Water Using Karl Fischer Reagent, ASTM E203-75.
      1990 Annual Book of ASTM Standards, Volume 15.05. Philadelphia, PA, 1990.

19.    Standard Practice for Sampling Industrial Chemicals, ASTM E300-86.  1990
      Annual Book of ASTM Standards, Volume 6.02.  Philadelphia, PA, 1990.

20.    Excerpts from Standard Practice for Use of the International System of Units
      (SI) (the Modernized Metric System), ASTM E380-89. 1990 Annual Book of
      ASTM Standards, Volume 6.01.  Philadelphia, PA, 1990.

21.    Standard  Practice for Conducting an Interlaboratory Study to Determine the
      Precision of a  Test Method, ASTM E691-87.  1990 Annual Book of ASTM
      Standards, Volume 6.01.  Philadelphia, PA, 1990.
Page 70

-------
           22.    Standard Guide for ASTM Standard Specification Quality Statements, ASTM
                 E1267-88.  1990 Annual Book of ASTM Standards, Volume 6.01. Philadelphia,
                 PA, 1990.

           Other Literature/Bibliographical References
           23.    Federal Test Method Standard No. 141c, Paint, Varnish, Lacquer, and Related
                 Materials; Methods of Inspection, Sampling, and Testing.  General Services
                 Administration, January 24, 1986.

           24.    EPA-340/1-84-001 a, VOC Sampling and Analysis Workshop, Volume I VOC
                 Reference Methods.  U.S. EPA, OAQPS, SSCD, Washington, DC, September
                 1983.

           25.    EPA-340/1-86-016, A Guideline for Surface Coating Calculations. U.S. EPA,
                 OAQPS, SSCD, Washington, DC, July 1986.

           26.    EPA 340/1-88-003, Recordkeeping Guidance  Document for Surface Coating
                 Operations and the Graphics Arts  Industry, U.S. EPA, SSCD, Washington, DC,
                 May 1989.

           27.    EPA-450/2-77-008, Control of Volatile Organic Emissions from  Existing Station-
                 ary Sources - Volume II: Surface Coating of Cans, Coils,  Paper, Fabrics,
                 Automobiles, and Light-Duty Trucks.  U.S. EPA, OAQPS,  Research  Triangle
                 Park, NC, May 1977.

           28.    EPA-450/3-84-019, Procedures for Certifying Quantity of Volatile Organic
                 Compounds Emitted by Paint,  Ink,  and Other  Coatings. U.S. EPA,  OAQPS,
                 ESED, Research Triangle Park, NC, December 1984.

           29.    EPA-600/4-77-027b, Quality Assurance Handbook for Air Pollution
                 Measurement Systems: Volume III. Stationary Source Specific Methods.  U.S.
                 EPA, EMSL, Research Triangle Park,  NC, November  1976.

           30.    EPA-600/9-76-005, Quality Assurance Handbook for Air Pollution Measurement
                 Systems: Volume I. Principles.  U.S. EPA, EMSL, Research Triangle Park, NC,
                 December 1984.

^          31.    Code of Federal  Regulations, General Provisions, Performance Tests, 40 CFR
                 60.8 Subpart A, Washington, DC, 1989.
*,

f
           32.    Code of Federal Regulations, General Provisions, Incorporation by Reference,
                 40 CFR 60.17 Subpart A, Washington,  DC, 1989.

                                                                               Page 71

-------
33.    Standards of Performance for Surface Coating of Metal Furniture, 40 CFR 60
      Subpart EE, Washington, DC, 1989.

34.    Standards of Performance for Automobile and Light-Duty Truck Surface Coating
      Operations, 40 CFR 60 Subpart MM, Washington, DC, 1989.

35.    Standards of Performance for the Graphic Arts Industry: Publication
      Rotogravure Printing, 40 CFR 60 Subpart QQ, Washington, DC,  1989.

36.    Standards of Performance for Pressure Sensitive Tape and Label Surface
      Coating Operations, 40 CFR 60 Subpart RR, Washington, DC, 1989.

37.    Standards of Performance for Industrial Surface Coating: Large Appliances, 40
      CFR 60 Subpart SS, Washington, DC, 1989.

38.    Standards of Performance for Metal Coil Surface Coating, 40 CFR 60 Subpart
      TT, Washington, DC, 1989.

39.    Standards of Performance for the Beverage Can Surface Coating Industry, 40
      CFR 60 Subpart WW, Washington,  DC, 1989.

40.    Standards of Performance for the Rubber Tire Manufacturing Industry, 40 CFR
      60 Subpart BBB, Washington, DC,  1989.

41.    Standards of Performance for Flexible Vinyl and Urethane Coating and Printing,
      40 CFR 60 Subpart FFF, Washington, DC, 1989.

42.    Standards of Performance for Magnetic Tape Coating Facilities,  40 CFR 60
      Subpart SSS, Washington, DC,  1989.

43.    Standards of Performance for Industrial Surface Coating: Surface Coating of
      Plastic Parts for Business Machines, 40 CFR 60 Subpart TTT, Washington, DC,
      1989.

44.    Code of Federal Regulations, Reference Methods 24 AND  24A, 40 CFR 60,
      Appendix A, Washington, DC 1989.

45.    Manual on Determination of Volatile Organic Compounds in Paints,  Inks, and
      Related Coating Products, J. John  Brezinski, ed., ASTM Manual Series: MNL 4,
      1989.
Page 72

-------
I
                             (Pit
   TECHNICAL REPORT DATA
read Imovcnons on me reverse Of fore comnietmii
 1. REPORT NQ.

   EPA-340/1-91-011
                            3. RECIPIENT'S ACCESSION NO.
4. TITLE ANO SUBTITLE
  Standard  Operating  Procedure for Analysis of Coating and
  Ink  Samples  by  Reference Methods 24 a.nd 24A
                            5. REPORT DATE

                             Seotember
                            ft. PERFORMING ORGANIZATION CODE
7. AUTHQR(S)
  Bruce A.  Olson, Melinda  K.  Wood, John T. Chehaske
                            3. PERFORMING ORGANIZATION REPORT NO.

                                  91-133-T4/S
9»PEflFORMINO ORGANIZATION NAME ANO AOQRESS
* Pacific Environmental Services, Inc.
'"SoO Herndon Parkway, Suite 200
Herndon, Virginia 22070-5225
10. PROGRAM ELEMENT NO.
WA 91-133
11 CONTRACT/GRANT NO.
68-02-4464
 12. SPONSORING AGENCY NAME ANO ADDRESS
   U.S.  Environmental  Protection Agency
   Office  of Air Quality  Planning and Standards
   Stationary  Source  Compliance Division
   Washington, D.C.   20460
                            13. TYPE OF REPORT ANO PERIOD COVERED
                                  Final	__
                            14. SPONSORING AGENCY CODE
 15. SUPPLEMENTARY NOTES
  SSCD, Organic Chemicals  Section, 401
  Washington, D.C.   20460  Phone:  (703)
      M St., S.W.
      308-3663
         See  a  so:
;EN-34W1) EPA/1-91-010:  Sampling
          EPA/1-91-012:  References
 18. ABSTRACT
       Standard procedures  are presented in this report for the handling and physical/
  chemical characterization of coating and ink samples that require analysis by EPA
  Reference Methods  24  or  24A (RM 24/24A) as found in 40 CFR 60, Appendix A.  For the
  purpose  of developing consistency among analytical laboratories nationwide, it
  delineates the activities deemed proper and necessary to ensure that an accurate and
  precise  measurement of coating  or ink sample properties are made.   If the activities  are
  conducted as described,  questions concerning the validity and reproducibility of analyses
  performed can be avoided.

         This Standard  Operating  Procedure (SOP) covers all types of  industrial coating
  and printing ink operations, regardless of the method of application of coating or ink,
  including, for example,  dip, spray,  roll, flow, electrostatic, or electro-deposition
  processes.  Special cases involving  use of exempt solvents dichloromethane (methylene
  chloride) and 1,1,1-trichloroethane  (methyl  chloroform) or multi-component mixture
  coating  formulations  are  covered in  accordance with most recent technical and regulatory
  thinking, however, this  is not  a regulatory but a technical procedure guideline document.
 17.
                                  KEY WORDS ANO DOCUMENT ANALYSIS
a. DESCRIPTORS
w Air Pollution Sampling
v Coatings VOC
Inks ASTM Methods
N Analysis
Solvents
18. OISTRI8UTION STATEMENT
b.lOENTIFIERS/OPGN £NOED TERMS
Analysis of Coatings
and Inks
19. SECURITY CLASS / Tins fltporll
Unclassified
20. SECURITY CLASS iTlut parr;
Unclassified
c. COSATi Field Group

73
22. PRICE j
 EPA f**m 2220-1 (H«». 4—77)    »<*cvious  COITION is OMOUCTK

-------