ENVIRONMENTAL PROTECTION AGENCY
PERFORMANCE TEST METHODS
U.S. ENVIRONMENTAL PROTECTION AGENCY
OFFICE OF ENFORCEMENT
OFFICE OF GENERAL ENFORCEMENT
WASHINGTON, D.C. 20460

-------

-------
                                      EPA-340/1 -78-011
ENVIRONMENTAL PROTECTION AGENCY
     PERFORMANCE TEST METHODS
                  Parts I and  II
                         by

                  PEDCo Environmental, Inc.
                    11499 Chester Road
                    Cincinnati, Ohio 45246
                      Prepared for

             U.S. ENVIRONMENTAL PROTECTION AGENCY
               Division of Stationary Source Enforcement
                    Office of Enforcement
                    Washington, DC 20460

                      August 1, 1978

-------
             STATIONARY SOURCE ENFORCEMENT SERIES

The Stationary Source Enforcement series of reports is issued by the Office
of General Enforcement, Environmental Protection Agency,  to assist the
Regional Offices in activities related to enforcement of implementation
plans, new source emission standards, and hazardous emission standards to
be developed under the Clean Air Act.  Copies of Stationary Source
Enforcement reports are available - as supplies permit - from the U.S.
Environmental Protection Agency, Office of Administration, Library
Services, MD-35, Research Triangle Park, North  Carolina 27711,  or may be
obtained, for a nominal cost, from the  National Technical Information
Service, 5285 Port Royal Road, Springfield,  Virginia 22161.
                               11

-------
                           FOREWORD





     This packet is part of a series of emission testing




related material that has been supplied to Federal, State




and local agencies by the Division of Stationary Source




Enforcement of the Environmental Protection Agency to enable




the field enforcement staff to keep informed of current




methods development and publications.




     Reference Methods 1 through 8 of the New Source




Performance Standards were revised and republished in the




Federal Register on August 18, 1977 with subsequent revisions




published on January 10, 1978.  However, due to the nature




of the Federal Register, print size and figures must be




reduced and many procedures referenced or implied in an




effort to reduce the number of pages.  This reduced size




makes it extremely difficult to set up and continually use




the actual Federal Register pages as a field test manual.




     To enable the field enforcement staff to develop and




maintain a current field testing manual, an updated, double




spaced, typed version of Reference Methods 1 through 8




(Part 1) and field data sheets (Part II) are provided in




this packet.  Additional material will be provided as it is




developed and published.

-------

-------
                      TABLE OF CONTENTS


                                                           Page

PART I - REFERENCE METHODS

Method 1 - Sample and Velocity Traverses for Stationary    1-1
           Sources

Method 2 - Determination of Stack Gas Velocity and Volu-   1-13
           metric Flow Rate

Method 3 - Gas Analysis for Carbon Dioxide, Excess Air,    1-45
           and Dry Molecular Weight

Method 4 - Determination of Moisture in Stack Gases        1-61

Method 5 - Determination of Particulate Emissions from     1-79
           Stationary Sources

Method 6 - Determination of Sulfur Dioxide Emissions       1-119
           from Stationary Sources

Method 7 - Determination of Nitrogen Oxide Emissions       1-135
           from Stationary Sources

Method 8 - Determination of Sulfuric Acid Mist and Sulfur  1-151
           Dioxide Emissions from Stationary Sources

PART II - EMISSION TESTING DATA SHEETS

Sample Point Location                                      II-l

Gas Velocity and Cyclonic Flow Determination               II-2

Pitot Tube Calibration                                     II-3

Integrated Bag Sample                                      II-4

Orsat Analysis                                             II-5

Method 4 - Field Sample and Recovery                       II-6

Method 5 - Field Sample                                    II-7

Method 5 - Recovery and Integrity                          II-8

Method 5 - Blank Analysis                                  II-9

Method 5 - Sample Analysis                                 11-10

-------
                TABLE OF CONTENTS (Continued)






                                                           Page



Nozzle Calibration                                         11-11




Dry Gas Meter Calibration                                  11-12




Dry Gas Meter Post-Test Check                              11-13




Stack Temperature Sensor Calibration                       11-14




Method 6 - Field Sample                                    11-15




Method 6 - Recovery and Integrity                          11-16




Method 6 - Sample Analysis                                 11-17




Method 7 - Field Sample                                    11-18




Method 7 - Recovery and Integrity                          11-19




Method 7 - Sample Analysis                                 11-20




Spectrophotometer Optimum Wave Length                      11-21




Method 8 - Field Sample                                    11-22




Method 8 - Recovery and Integrity                          11-23




Method 8 - Sample Analysis                                 11-24




Example Sample Label                                       11-25

-------
REFERENCE METHODS



   1 THROUGH 8







     PART 1

-------

-------
SAMPLE AND VELOCITY TRAVERSES



   FOR STATIONARY SOURCES








          METHOD 1

-------

-------
                             1-1
         METHOD 1—SAMPLE AND VELOCITY TRAVERSES
                 FOR STATIONARY SOURCES
1.  Principle and Applicability
     1.1  Principle.  To aid in the representative measurement of
pollutant emissions and/or total volumetric flow rate from a
stationary source, a measurement site where the effluent stream
is flowing in a known direction is selected, and the cross-section
of the stack is divided into a number of equal areas.  A traverse
point is then located within each of these equal areas.
     1.2  Applicability.  This method is applicable to flowing gas
streams in ducts, stacks, and flues.  The method cannot be used when:
(1) flow is cyclonic or swirling (see Section 2.4), (2) a stack is
                                                              o
smaller than about 0.30 meter (12 in.) in diameter, or 0.071 m
        2
(113 in. ) in cross-sectional area, or (3) the measurement site is less
than two stack or duct diameters downstream or less than a half
diameter upstream from a flow disturbance.
     The requirements of this method must be considered before con-
struction of a new facility from which emissions will be measured;
failure to do so may require subsequent alterations to the stack or
deviation from the standard procedure.  Cases involving variants are
subject to approval by the Administrator, U.S. Environmental Protection
Agency.
2.  Procedure
     2.1  Selection of Measurement Site.   Sampling or velocity measure-
ment is performed at a site located at least eight stack or duct

-------
                             1-2
diameters downstream and two diameters upstream from any flow
disturbance such as a bend, expansion, or contraction in the stack,
or from a visible flame.  If necessary, an alternative location  may
be selected, at a position at least two stack  or duct diameters
downstream and a half diameter upstream from any flow disturbance.
For a rectangular cross section, an equivalent diameter (D  ) shall
be calculated from the following equation, to  determine the upstream
and downstream distances:

          D   -   2LW
          'e     L + W
where L = length and W = width.
     2.2  Determining the Number of Traverse Points.
     2.2.1  Particulate Traverses.   When the eight-  and  two-diameter
criterion can be met, the minimum number of traverse points  shall  be:
(1) twelve, for circular or rectangular stacks with  diameters  (or
equivalent diameters) greater than 0.61 meter (24 in.);  (2)  eight,
for circular stacks with diameters between 0.30 and  0.61  meter
(12-24 in.); and (3) nine, for rectangular stacks with equivalent
diameters between 0.30 and 0.61  meter (12-24 in.).
     When the eight- and two-diameter criterion cannot be met, the
minimum number of traverse points is determined from Figure  1-1.
Before referring to the figure,  however, determine the distances  from
the chosen measurement site to the nearest upstream and  downstream
disturbances, and divide each distance by the stack  diameter or
equivalent diameter, to determine the distance in terms  of the number
of duct diameters.  Then, determine from Figure 1-1  the  minimum number

-------
                             1-3
                                                              — &)   CD
UJ
u
UJ
o
z
<
00
cc
CO
5

I
u.

o
cc
LL
CO
a.
UJ
5
o

a
a>
in

0)

a:
                                                                         JD

                                                                          D
                                                                          U
ro
a.
i/)
4-*
C

'o
Q.
 a>

 ra
 D
 C

 E
 3

 E
 en
                 siNiod 3sa3Avai jo H38wni\i

-------
                              1-4
of traverse points that corresponds:   (1)  to the number of duct
diameters upstream; and (2) to the number  of diameters downstream.
Select the higher of the two minimum numbers of traverse points, or
a greater value, so that for circular stacks the number is a
multiple of 4, and for rectangular stacks, the number is one of those
shown in Table 1-1.
   Table 1-1.  CROSS-SECTIONAL LAYOUT FOR  RECTANGULAR STACKS
               No. of traverse          Matrix
                   points               layout
                     9                   3x3
                    12                   4x3
                    16                   4x4
                    20                   5x4
                    25                   5x5
                    30                   6x5
                    36                   6x6
                    42                   7x6
                    49                   7x7
     2.2.2  Velocity (Non-Particulate) Traverses.  When velocity or
volumetric flow rate is to be determined (but not particulate matter),
the same procedure as that for particulate traverses (Section 2.2.1)
is followed, except that Figure 1-2 may be used instead of Figure 1-1.
     2.3  Cross-Sectional Layout and Location of Traverse Points.
     2.3.1  Circular Stacks.  Locate the traverse points on two perpendi-
cular diameters according to Table 1-2 and the example shown in
Figure 1-3.  Any equation (for examples, see Citations 2 and 3 in the

-------
                                     1-5
     in
     CM
     o

     1M
UU
O
co

Q
O


<
CO
cc
CO

Q


O
O
CC
cc
f-
co
Q.
D
co   9
oc   «-
LU
o

o
              00
              oc
                                                                                O)

                                                                                03
                                                                         00
                                                                         LU
                                                                         u
                                                                         z
                                                                         <

                                                                         CO

                                                                     00  §
                                                                         LU
                                                                         O
                                                                         z
                                                                         <
                                                                         00
                                                                         cc
                                                                         00
                                                                         Q


                                                                         o
                                                                     ID  U|J


                                                                         o
                                                                         cc
                                                                         «f
                                                                      m  S
                                                                         cc
                                                                         K
                                                                         co
                                                                         Z

                                                                         O

                                                                      -  2
                                                                         cc
                                                                         LU
                                                                         t-
                                                                         LU
                                                                         5

                                                                      ro  Q
                                                                         Q
                                                                      (N
                                                                        (T3
                                                                        Q.
                                                                        c
                                                                        o
                                                                        c
                                                                         >
                                                                         L_
                                                                        ^O

                                                                         in
                                                                         *-•
                                                                         C

                                                                        'o
                                                                         Q.

                                                                         
-------
                          1-6
TRAVERSE
POINT
1
2
3
4
5
6
DISTANCE.
% of diameter
4.4
14.7
29.5
70.5
85.3
95.6
Figure 1-3. Example showing circular stack cross section divided into
12 equal areas, with location of traverse points indicated,

-------
                             1-7
Table 1-2. LOCATION OF TRAVERSE POINTS IN CIRCULAR STACKS
      (Percent of stack diameter from inside wall to traverse point)
Traverse
point
number
on a
diameter
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
Number of traverse points on a diameter
2
14.6
85.4






















4
6.7
25.0
75.0
93.3




















6
4.4
14.6
29.6
70.4
85.4
95.6


















8
3.2
10.5
19.4
32.3
67.7
80.6
89.5
96.8
















10
2.6
8.2
14.6
22.6
34.2
65.8
77.4
85.4
91.8
97.4














12
2.1
6.7
11.8
17.7
25.0
35.6
64.4
75.0
82.3
88.2
93.3
97.9












14
1.8
5.7
9.9
14.6
20.1
26.9
36.6
63.4
73.1
79.9
85.4
90.1
94.3
98.2










16
1.6
4.9
8.5
12.5
16.9
22.0
28.3
37.5
62.5
71.7
78.0
83.1
87.5
91.5
95.1
98.4








18
1.4
4.4
7.5
10.9
14.6
18.8
23.6
29.6
38.2
61.8
70.4
76.4
81.2
85.4
89.1
92.5
95.6
98.6






20
1.3
3.9
6.7
9.7
12.9
16.5
20.4
25.0
30.6
38.8
61.2
69.4
75.0
79.6
83.5
87.1
90.3
93.3
96.1
98.7




22
1.1
3.5
6.0
8.7
11.6
14.6
18.0
21.8
26.2
31.5
39.3
60.7
68.5
73.8
78.2
82.0
85.4
88.4
91.3
94.0
96.5
98.9


24
1.1
3.2
5.5
7.9
10.5
13.2
16.1
19.4
23.0
27.2
32.3
39.8
60.2
67.7
7:. 8
77.0
80.6
83.9
86.8
89.5
92.1
94.5
96.8
98.9

-------
                               1-8
Bibliography) that gives the same values as those in Table 1-2 may
be used in lieu of Table 1-2.
     For particulate traverses, one  of the diameters must be in a plane
containing the greatest expected concentration variation, e.g., after
bends, one diameter shall  be in the  plane of the bend.   This requirement
becomes less critical as the distance from the disturbance increases;
therefore, other diameter locations  may be used, subject to approval  of
the Administrator.
     In addition, for stacks having  diameters greater than 0.61 m
(24 in.) no traverse points shall be located within 2.5 centimeters
(1.00 in.) of the stack walls;  and for stack diameters  equal to or less
than 0.61  m (24 in.), no traverse points shall be located within 1.3  cm
(0.50 in.) of the stack walls.   To meet these criteria, observe the
procedures given below.
     2.3.1.1  Stacks With Diameters  Greater Than 0.61 m (24 in.). When
any of the traverse points as located in Section 2.3.1  fall within
2.5 cm (1.00 in.) of the stack  walls, relocate them away from the
stack walls to:  (1) a distance of 2.5 cm (1.00 in.); or (2) a distance
equal to the nozzle inside diameter, whichever is larger.  These
relocated traverse points (on each end of a diameter) shall be the
"adjusted" traverse points.
     Whenever two successive traverse points are combined to form a
single adjusted traverse point, treat the adjusted point as two
separate traverse points, both  in the sampling (or velocity measure-
ment) procedure, and in recording the data.

-------
                               1-9
     2.3.1.2  Stacks With Diameters Equal to or Less Than 0.61 m (24
in.).  Follow the procedure in Section 2.3.1.1, noting only that any
"adjusted" points should be relocated away from the stack walls to:
(1) a distance of 1.3 cm (0.50 in.); or (2) a distance equal to the
nozzle inside diameter, whichever is larger.
     2.3.2  Rectangular Stacks.  Determine the number of traverse points
as explained in Sections 2.1 and 2.2 of this method.  From Table 1-1,
determine the grid configuration.  Divide the stack cross-section into as
many equal rectangular elemental areas as traverse points, and then
locate a traverse point at the centroid of each equal  area according to
the example in Figure 1-4.
     If the tester desires  to use more than the minimum number of
traverse points, expand the "minimum number of traverse points" matrix
(see Table 1-1) by adding the extra traverse points along one or the
other or both legs of the  matrix; the final matrix need not be balanced.
For example, if a 4 x 3 "minimum number of points" matrix were expanded
to 36 points, the final matrix could be 9 x 4 or 12 x  3, and would not
necessarily have to be 6 x  6.   After constructing the  final  matrix,
divide the stack cross-section into as many equal rectangular, elemental
areas as traverse points, and locate a traverse point  at the centroid of
each equal area.
     The situation of traverse points being too close  to the stack walls
is not expected to arise with rectangular stacks.  If  this problem
should ever arise, the Administrator must be contacted for resolution of
the matter.

-------
                  1-10
                          o    I   o
                               I
                          —-r--
                          o    I   o
                               I
Figure 1-4. Example showing rectangular stack cross
section divided into 12 equal areas, with a traverse
point at centroid of each area.

-------
                                i-n
     2.4  Verification of Absence of Cyclonic Flow.  In most stationary
sources, the direction of stack gas flow is essentially parallel to the
stack walls.  However, cyclonic flow may exist (1) after such devices as
cyclones and inertial demisters following venturi scrubbers, or (2) in
stacks having tangential inlets or other duct configurations which tend
to induce swirling; in these instances, the presence or absence of
cyclonic flow at the sampling location must be determined. The following
techniques are acceptable for this determination.
     Level and zero the manometer.  Connect a Type S pi tot tube to the
manometer.  Position the Type S pi tot tube at each traverse point, in
succession, so that the planes of the face openings of the pitot tube
are perpendicular to the stack cross-sectional plane: when the Type S
pitot tube is in this position, it is at "0° reference."  Note the
differential pressure (Ap) reading at each traverse point.  If a null
(zero) pitot reading is obtained at 0° reference at a given traverse
point, an acceptable flow condition exists at that point.   If the pitot
reading is not zero at 0° reference, rotate the pitot tube (up to +90°
yaw angle), until a null reading is obtained.  Carefully determine and
record the value of the rotation angle (a) to the nearest degree.   After
the null technique has been applied at each traverse point, calculate
the average of the absolute values of a; assign a values of 0° to those
points for which no rotation was required, and include these in the
overall average.  If the average value of a is greater than 10°, the
overall flow condition in the stack is unacceptable and alternative
methodology, subject to the approval of the Administrator, must be used
to perform accurate sample and velocity traverses.

-------
                                1-12
3.  Bibliography
     1.  Determining Dust Concentration in a Gas Stream.   ASME.  Per-
formance Test Code No. 27.  New York.   1957.
     2.  Devorkin, Howard, et al.   Air Pollution Source Testing  Manual.
Air Pollution Control District.  Los Angeles, CA.   November 1963.
     3.  Methods for Determination of Velocity,  Volume, Dust and Mist
Content of Gases.  Western Precipitation Division  of Joy  Manufacturing
Co.  Los Angeles, CA.  Bulletin WP-50.  1968.
     4.  Standard Method for Sampling  Stacks for Particulate Matter. In:
1971 Book of ASTM Standards, Part  23.   ASTM Designation D-2928-71.
Philadelphia, PA.  1971.
     5.  Hanson, H. A., et al.   Particulate Sampling Strategies  for
Large Power Plants Including Nonuniform Flow. USEPA, ORD,  ESRL, Research
Triangle Park, N.C.  EPA-600/2-76-170.  June 1976.
     6.  Entropy Environmentalists, Inc.  Determination of the Optimum
Number of Sampling Points:  An  Analysis of Method  1  Criteria.  Environmental
Protection Agency.  Reserch Triangle Park, N. C. EPA Contract No.  68-01-3172,
Task 7.

-------
DETERMINATION OF STACK GAS VELOCITY AND VOLUMETRIC



          FLOW RATE (TYPE S PITOT TUBE)





                   METHOD 2

-------
                              1-13
    METHOD 2—DETERMINATION OF STACK GAS VELOCITY AND
         VOLUMETRIC FLOW RATE (TYPE S PITOT TUBE)
1.   Principle and Applicability
     1.1  Principle.  The average gas velocity in a stack is
determined from the gas density and from measurement of the average
velocity head with a Type S (Stausscheibe or reverse type) pi tot tube.
     1.2  Applicability.  This method is applicable for measurement
of the average velocity of a gas stream and for quantifying gas flow.
     This procedure is not applicable at measurement sites which fail
to meet the criteria of Method 1, Section 2.1.  Also, the method cannot
be used for direct measurement in cyclonic or swirling gas streams;
Section 2.4 of Method 1 shows how to determine cyclonic or swirling
flow conditions.  When unacceptable conditions exist, alternative
procedures, subject to the approval of the Administrator, U.  S.
Environmental Protection Agency, must be employed to make accurate
flow rate determinations; examples of such alternative procedures are:
(1) to install straightening vanes; (2) to calculate the total volumetric
flow rate stoichiometrically, or (3) to move to another measurement
site at which the flow is acceptable.
2.   Apparatus
     Specifications for the apparatus are given below.  Any other
apparatus that has been demonstrated (subject to approval of the
Administrator) to be capable of meeting the specifications will be
considered acceptable.
     2.1  Type S Pi tot Tube.  The Type S pi tot tube (Figure 2-1) shall
be made of metal tubing (e.g., stainless steel).  It is recommended
that the external tubing diameter (dimension D., Figure 2-2b) be

-------
                                         1-14
1.90-2.54 cm*
(0.75  1.0 in.)
       •   ,  7.62 cm (3 in.)'
                            TEMPERATURE SENSOR
                           7H5 £
                             TYPE SPITOT TUBE
           •SUGGESTED (INTERFERENCE FREE)
            PITOT TUBE THERMOCOUPLE SPACING
                                                          LEAK-FREE
                                                          CONNECTIONS
                                           MANOMETER
                      Figure 2-1.  Type S pitot tube manometer assembly.

-------
                                1-15
between 0.48 and 0.95 centimeters (3/16 and 3/8 inch).   There shall
be an equal distance from the base of each leg of the pi tot tube to
its face-opening plane (dimensions P^ and Pg, Figure 2-2b); it is
recommended that this distance be between 1.05 and 1.50 times the
external tubing diameter.  The face openings of the pitot tube shall,
preferably, be aligned as shown in Figure 2-2; however, slight mis-
alignments of the openings are permissible (see Figure 2-3).
     The Type S pitot tube shall have a known coefficient, determined
as outlined in Section 4.  An identification number shall be assigned
to the pitot tube; this number shall be permanently marked or
engraved on the body of the tube.
     A standard pitot tube may be used instead of a Type S, provided
that it meets the specifications of Sections 2.7 and 4.2; note, however,
that the static and impact pressure holes of standard pitot tubes are
susceptible to plugging in particulate-laden gas streams.  Therefore,
whenever a standard pitot tube is used to perform a traverse, adequate
proof must be furnished that the openings of the pitot tube have not
plugged up during the traverse period; this can be done by taking a
velocity head (AP) reading at the final traverse point, cleaning out
the impact and static holes of the standard pitot tube by "back-
purging" with pressurized air, and then taking another Ap reading.
If the Ap readings made before and after the air purge are the same
(+5 percent), the traverse is acceptable.  Otherwise, reject the run.
Note that if Ap at the final traverse point is unsuitably low,
another point may be selected.  If "back-purging" at regular intervals

-------
                        1-16
    TRANSVERSE
    TUBE AXIS
             \
                         FACE
                    h~-OPENING
                        PLANES
                         A SIDE PLANE
LONGITUDINAL '
TUBE AXIS *":
)
\
Dt
*
A
B
                                        PA
NOTE:

1.05 D
                        B SIDE PLANE

                          (b)
                        AORB
                          (c)
Figure 22. Properly constructed Type S pitot tube, shown
in:  (a) end view; face opening planes perpendicular to trans-
verse axis; (b) top view; face opening planes parallel to lon-
gitudinal axis; (c)  side view; both legs of equal length and
centerlines coincident, when viewed from both sides.  Base-
line coefficient values of 0.84 may be assigned to pitot tubes
constructed this way.

-------
        TRANSVERSE
         TUBE AXIS
                                           1-17
                                                                 (b)
LONGITUDINAL
  TUBE AXIS —
                                                   (e)
                                                                     '    J. 01 <+ or -)
                                                                     	j_LJT • ^	  IT —
                                                   (f)
                                                   (8)

             Figure 2-3.  Types of face-opening misalignment that can result from field use or im-
             proper construction of Type S pilot tubes.  These will not affect the baseline value
             of Cp(s) so long as a\ and 0.2 < 10°, 01 and 02 < 5°, z < 0.32 cm (1/8 in.) and w <
             0.08cm (1/32 in.) (citation 11  in Section 6).

-------
                                 1-18
is part of the procedure, then comparative Ap readings shall  be taken,
as above, for the last two back purges at which suitably high Ap
readings are observed.
     2.2  Differential Pressure Gauge.   An inclined manometer or
equivalent device is used.  Most sampling trains are equipped with
a 10-in. (water column) inclined-vertical manometer, having 0.01-ln.
HpO divisions on the 0- to 1-in. inclined scale, and 0.1-in.  HpO
divisions on the 1- to 10-in.  vertical  scale.  This type of manometer
(or other gauge of equivalent sensitivity) is satisfactory for the
measurement of Ap values as low as  1.3 mm (0.05 in.) HA0.   However,  a
differential pressure gauge of greater sensitivity shall  be used
(subject to the approval of the Administrator), if any of the following  is
found to be true:  (1) the arithmetic average of all Ap readings at
the traverse points in the stack is less than 1.3 mm (0.05 in.) hLO;
(2) for traverses of 12 or more points, more than 10 percent  of the
individual Ap readings are below 1.3 mm (0.05 in.) H^O; (3) for traverses
of fewer than 12 points, more  than  one  Ap reading is below 1.3 mm
(0.05 in.) HpO.  Citation 18 in Section 6 describes commercially
available instrumentation for  the measurement of low-range gas velocities.
     As an alternative to criteria  (1)  through (3) above,  the following
calculation may be performed to determine the necessity of using a more
sensitive differential pressure gauge:
                  n
                  I
          T  =            .
                  I    ^p~
                i = 1      ]
where:
     Ap. = Individual  velocity head reading  at  a  traverse  point,
           rrni H0 (1n. H0)

-------
                                1-19
     n   = Total number of traverse points
     K   = 0.13 mm H20 when metric units are used and 0.005 in. HpO
           when English units are used
If T is greater than 1.05, the velocity head data are unacceptable
and a more sensitive differential  pressure gauge must be used.
     Note:  If differential pressure gauges other than inclined
manometers are used (e.g., magnehelic gauges), their calibration
must be checked after each test series.  To check the calibration
of a differential pressure gauge, compare Ap readings of the gauge
with those of a gauge-oil manometer at a minimum of three points,
approximately representing the range of Ap values in the stack.  If,
at each point, the values of Ap as read by the differential pressure
gauge and gauge-oil manometer agree to within 5 percent, the differential
pressure gauge shall be considered to be in proper calibration.  Other-
wise, the test series shall either be voided, or procedures to adjust
the measured Ap values and final results shall be used, subject 'to the
approval of the Administrator.
     2.3  Temperature Gauge.  A thermocouple, liquid-filled bulb
thermometer, bimetallic thermometer, mercury-in-glass thermometer, or
other gauge capable of measuring temperature to within 1.5 percent of
the minimum absolute stack temperature shall be used.  The temperature
gauge shall be attached to the pi tot tube such that the sensor tip does not
touch any metal; the gauge shall be in an interference-free arrangement
with respect to the pitot tube face openings (see Figure 2-1 and also
Figure 2-7 in Section 4).  Alternate positions may be used if the pitot
tube-temperature gauge system is calibrated according to the procedure

-------
                                1-20
of Section 4,  Provided that a difference of not more than 1  percent
in the average velocity measurement is introduced, the temperature
gauge need not be attached to the pitot tube; this alternative is
subject to the approval of the Administrator.
     2.4  Pressure Probe and Gauge.  A piezometer tube and mercury-
or water-filled U-tube manometer capable of measuring stack pressure
to within 2.5 mm (0.1 in.) Hg is used.  The static tap of a standard
type pitot tube or one leg of a Type S pitot tube with the face opening
planes positioned parallel to the gas flow may also be used as the
pressure probe.
     2.5  Barometer.  A mercury, aneroid, or other barometer capable
of measuring atmospheric pressure to within 2.5 mm Hg (0.1 in. Hg) may
be used.  In many cases, the barometric reading may be obtained from
a nearby national weather service station, in which case the station
value (which is the absolute barometric pressure) shall be requested
and an adjustment for elevation differences between the weather station
and the sampling point shall be applied at a rate of minus 2.5 mm
(0.1 in.) Hg per 30 meter (100 foot) elevation increase, or vice-versa
for elevation decrease.
     2.6  Gas Density Determination Equipment.  Method 3 equipment, if
needed (see Section 3.6), to determine the stack gas dry molecular
weight, and Reference Method 4 or Method 5 equipment for moisture content
determination; other methods may be used subject to approval of the
Administrator.
     2.7  Calibration Pitot Tube.  When calibration of the Type S pitot
tube is necessary (see Section 4), a  standard pitot tube is used as a
reference.  The standard  pitot tube shall, preferably, have a known

-------
                                 1-21
coefficient, obtained either (1) directly from the National Bureau
of Standards, Route 70 S, Quince Orchard Road, Gaithersburg,
Maryland, or (2) by calibration against another standard pitot tube
with an NBS-traceable coefficient.  Alternatively, a standard pitot
tube designed according to the criteria given in 2.7.1 through 2.7.5
below and illustrated in Figure 2-4 (see also Citations 7,8, and 17
in Section 6) may be used.  Pitot tubes designed according to these
specifications will have baseline coefficients of about 0.99 ^0.01.
     2.7.1  Hemispherical (shown in Figure 2-4), ellipsoidal, or
conical tip.
     2.7.2  A minimum of six diameters straight run (based upon D, the
external diameter of the tube) between the tip and the static pressure
holes.
     2.7.3  A minimum of eight diameters straight run between the static
pressure holes and the center!ine of the external tube, following the
90 degree bend.
     2.7.4  Static pressure holes of equal size (approximately 0.1 D),
equally spaced in a piezometer ring configuration.
     2.7.5  Ninety degree bend, with curved or mitered junction.
     2.8  Differential Pressure Gauge for Type S Pitot Tube Calibration.
An inclined manometer or equivalent is used.   If the single-velocity
calibration technique is employed (see Section 4.1.2.3),  the calibration
differential pressure gauge shall be readable to the nearest 0.13 mm
H«0 (0.005 in.  HpO).  For multivelocity calibrations,  the gauge  shall  be
readable to the nearest 0.13 mm H20 (0.005 in H20)  for Ap values  between 1.3
and 25 mm H20 (0.05 and 1.0 in. H20), and to  the nearest  1.3 mm  HgO
(0.05 in. H20) for Ap values above 25 mm H20  (1.0 in.  H20).   A special,  more
sensitive gauge will be required to read Ap values  below  1.3 mm  H20  [0.05
in. H20](see Citation 18 in Section 6).

-------
             1-22
                                         O
                                         U
3 K
U UJ
                                         Q.
                                         
-------
                               1-23
3.  Procedure
     3.1  Set up the apparatus as shown in Figure 2-1.   Capillary
tubing or surge tanks installed between the manometer and pitot tube
may be used to dampen Ap fluctuations.  It is recommended, but not
required, that a pretest leak-check be conducted, as follows:   (1) blow
through the pitot impact opening until at least 7.6 cm (3 in.)
HpO velocity pressure registers on the manometer; then, close  off the
impact opening.  The pressure shall remain stable for at least 15
seconds; (2) do the same for the static .pressure side, except  using
suction to obtain the minimum of 7.6 cm (3 in.) H?0.  Other leak-check
procedures, subject to the approval of the Administrator, may  be used.
     3.2  Level and zero the manometer.  Because the manometer level
and zero may drift due to vibrations and temperature changes,  make
periodic checks during the traverse.  Record all necessary data as
shown in the example data sheet (Figure 2-5).
     3.3  Measure the velocity head and temperature at the traverse
points specified by Method 1.  Ensure that the proper differential
pressure gauge is being used for the range of Ap values encountered
(see Section 2.2).  If it is necessary to change to a more sensitive
gauge, do so, and remeasure the Ap and temperature readings at each
traverse point.  Conduct a post-test leak-check (mandatory), as
described in Section 3.1 above, to validate the traverse run.
     3.4  Measure the static pressure in the stack.  One reading is
usually adequate.
     3.5  Determine the atmospheric pressure.
     3.6  Determine the stack gas dry molecular weight.  For combustion
processes or processes that emit essentially C02, 02» CO, and  N2> use

-------
                                      1-24
PLANT
DATE .
         RUN NO.
STACK DIAMETER OR DIMENSIONS, m(m.)
BAROMETRIC PRESSURE, mm Hg (in. Hg)	
CROSS SECTIONAL AREA, m2(ft2)	
OPERATORS 	
PITOTTUBEI.D.NO.
  AVG. COEFFICIENT, Cp =
  LAST DATE CALIBRATED.
                                      SCHEMATIC OF STACK
                                         CROSS SECTION
   Traverse
    Pt. No.
 Vel. Hd.,Ap
mm (in.) H20
                                  Stack Temperature
ts, °C
mm Hg (in.Hg)
                                  Average
                     Figure 2-5. Velocity traverse data.

-------
                                 1-25
Method 3.  For processes emitting essentially air, an analysis need
not be conducted; use a dry molecular weight of 29.0.  For other processes,
other methods, subject to the approval of the Administrator, must be used.
     3.7  Obtain the moisture content from Reference Method 4 (or
equivalent.) or from Method 5.
     3.8  Determine the cross-sectional area of the stack or duct at
the sampling location.  Whenever possible, physically measure the stack
dimensions rather than using blueprints.
4.  Calibration
     4.1  Type S  Pi tot Tube.  Before its initial use, carefully examine
the Type S pitot tube in top, side, and end views to verify that the face
openings of the tube are aligned within the specifications illustrated in
Figure 2-2 or 2-3.  The pitot tube shall not be used if it fails to meet
these alignment specifications.
     After verifying the face opening alignment, measure and record the
following dimensions of the pitot tube: (a) the external tubing diameter
(dimension D., Figure 2-2b); and (b) the base-to-opening plane distances
(dimensions P. and Pg, Figure 2-2b).  If D. is between 0.48 and 0.95 cm
(3/16 and 3/8 in.), and if P. and Pg are equal and between 1.05 and 1.50
D., there are two possible options:  (1) the pitot tube may be cali-
brated according to the procedure outlined in Sections 4.1.2 through
4.1.5 below, or (2) a baseline (isolated tube) coefficient value of
0.84 may be assigned to the pitot tube.  Note, however, that if the
pitot tube is part of an assembly, calibration may still be required,
despite knowledge of the baseline coefficient value  (see Section 4.1.1).
     If Dt, PA, and PB are outside the specified limits, the pitot tube
must be calibrated as outlined in 4.1.2 through 4.1.5 below.

-------
                               1-26
     4.1.1   Type S Pitot Tube Assemblies.   During sample and velocity
traverses,  the isolated Type S pi tot  tube  is not always used; in many
instances,  the pi tot tube is used  in  combination with other source-
sampling components (thermocouple, sampling probe, nozzle)  as part
of an "assembly."  The presence of other sampling components can
sometimes affect the baseline value of the Type S pitot tube
coefficient (Citation 9 in Section 6); therefore an assigned (or
otherwise known) baseline coefficient value may or may not  be valid
for a given assembly.  The baseline an^ assembly coefficient values
will be identical only when the relative placement of the components
in the assembly is such that aerodynamic interference effects are
eliminated.  Figures 2-6 through 2-8 illustrate interference-free
component arrangements for Type S pitot tubes having external tubing
diameters between 0.48 and 0.95 cm (3/16 and 3/8 in.).   Type S pitot
tube assemblies that fail to meet any or all of the specifications of
Figures 2-6 through 2-8 shall be calibrated according to the procedure
outlined in Sections 4.1.2 through 4.1.5 below, and prior to calibra-
tion, the values of the intercomponent spacings (pitot-nozzle, pitot-
thermocouple, pitot-probe sheath) shall be measured and recorded.
     Note:   Do not use any Type S pitot tube assembly which is
constructed such that the impact pressure opening plane of  the pitot
tube is below the entry plane of the nozzle (see Figure 2-6b).
     4.1.2  Calibration Setup.  If the Type S pitot tube is to be
calibrated, one leg of the tube shall be permanently marked A, and the
other, B.  Calibration shall be done in a flow system having the follow-
ing essential design features:
     4.1.2.1  The flowing gas stream must be confined to a  duct of
definite cross-sectional area, either circular or rectangular.  For

-------
                                    1-27
                           TYPE SPITOT TUBE
                          i > 1.90 cm (3/4 in.) FOR Dn - 1.3 cm (1/2 in.)
              SAMPLING NOZZLE
        A.  BOTTOM VIEW; SHOWING MINIMUM PITOT-NOZZLE SEPARATION.
SAMPLING
 PROBE
\
                    SAMPLING
                     NOZZLE
               TYPES
              PITOTTUBE
                              NOZZLE ENTRY
                                 PLANE
          a  SIDE VIEW; TO PREVENT PITOTTUBE
             FROM INTERFERING WITH GAS FLOW
             STREAMLINES APPROACHING THE
             NOZZLE. THE IMPACT PRESSURE
             OPENING PLANE OF THE PITOT TUBE
             SHALL BE EVEN WITH OR ABOVE THE
             NOZZLE ENTRY PLANE.
STATIC PRESSURE
 OPENING PLANE
                                                               IMPACT PRESSURE
                                                                OPENING PLANE
      Figure 2-6.  Proper pitot tube - sampling nozzle configuration to prevent
      aerodynamic interference; buttonhook - type nozzle; centers of nozzle
      and pitot opening aligned; Dt between 0.48 and 0.95 cm (3/16 and
      3/8 in.).

-------
         1-28
u
o

ff
        V)

        LU
                                      O)
                                      o
£)

 u.
 O>
 4-»
 C
                                      2.E

                                      c9?
                                      a> co
  ro
E '
u
eg
<0
p«'
A\
$ ,




I
c
C


u
•s-
A
rsi
«— -
^

LU
-J
i*
THERMOCOUI



;



UJ
ea
3
»-
»-
O
J-
Z
V)
UJ
a.
>
a
^
                                      u  u
                                      a> O
                                      aoo
                                       .  c
                                     i>.  QJ

                                     c^|
                                      a)  *-
                                      C.   «-
                                     U-Q

-------
 1-29
LO
o-
           u
          CM
          to
          ^.

          A\
  f }
  ^
                     OJ
                     o

                     O!
                     IV
                     ^
                     c
                     c
                     cu
a
o
4-i

TJ
a>
TJ
a>
QJ
C
C
o
                     o c
                     ~ ro
                    Is
                    is
                     OJ r-
                     tf> C
                     E?
                     3 C
                     e «
                     .  c
                    c»  a>
                     nj ^^
                     u V
                     3 Jd
                    .21 *-
                    a.Q

-------
                                 1-30
circular cross-sections, the minimum duct diameter shall  be 30.5 cm
(12 in.); for rectangular cross-sections, the width (shorter side)
shall be at least 25.4 cm (10 in.).
     4.1.2.2  The cross-sectional area of the calibration duct must
be constant over a distance of 10 or more duct diameters.  For a
rectangular cross-section, use an equivalent diameter, calculated
from the following equation, to determine the number of duct diameters:
where:
          De  =  TI                                   Equation 2-1
     D  = Equivalent diameter
     L  = Length
     W  = Width
     To ensure the presence of stable, fully developed flow patterns
at the calibration site, or "test section," the site must be located
at least eight diameters downstream and two diameters upstream from
the nearest disturbances.
     Note:  The eight- and two-diameter criteria are not absolute;
other test section locations may be used (subject to approval of the
Administrator), provided that the flow at the test site is stable and
demonstrably parallel to the duct axis.
     4.1.2.3  The flow system shall have the capacity to generate a
test-section velocity around 915 m/min (3000 ft/min).  This velocity
must be constant with time to guarantee steady flow during calibration.
Note that Type S pitot tube coefficients obtained by single-velocity
calibration at 915 m/min (3000 ft/min) will generally be valid to
within +3 percent for the measurement of velocities above 305 m/min

-------
                                  1-31
(1000 ft/min) and to within +5 to 6 percent for the measurement of
velocities between 180 and 305 m/min (600 and 1000 ft/min).  If a
more precise correlation between C  and velocity is desired, the flow
system shall have the capacity to generate at least four distinct,
time-invariant test-section velocities covering the velocity range
from 180 to 1525 m/min (600 to 5000 ft/min), and calibration data
shall be taken at regular velocity intervals over this range (see
Citations 9 and 14 in Section 6 for details).
     4.1.2.4  Two entry ports, one each for the standard and Type S
pitot tubes, shall be cut in the test section; the standard pitot entry
port shall be located slightly downstream of the Type S port, so that
the standard and Type S impact openings will lie in the same cross-
sectional plane during calibration.  To facilitate alignment of the
pitot tubes during calibration, it is advisable that the test section
be constructed of plexiglas or some other transparent material.
     4.1.3  Calibration Procedure.  Note that this procedure is a
general one and must not be used without first referring to the special
considerations presented in Section 4.1.5.  Note also that this pro-
cedure applies only to single-velocity calibration.  To obtain calibration
data for the A and B sides of the Type S pitot tube, proceed as follows:
     4.1.3.1  Make sure that the manometer is properly filled and that
the oil is free from contamination and is of the proper density.  Inspect
and leak-check all pitot lines; repair or replace if necessary.
     4.1.3.2  Level and zero the manometer.  Turn on the fan and allow
the flow to stabilize.  Seal the Type S entry port.
     4.1.3.3  Ensure that the manometer is level and zeroed.  Position
the standard pitot tube at the calibration point (determined as outlined

-------
                                 1-32
in Section 4.1.5.1), and align the tube so that its tip is pointed
directly into the flow.  Particular care should be taken in aligning
the tube to avoid yaw and pitch angles.  Make sure that the entry
port surrounding the tube is properly sealed.
     4.1.3.4  Read Apstd and record its value in a data table similar
to the one shown in Figure 2-9.  Remove the standard pitot tube from
the duct and disconnect it from the manometer.  Seal the standard
entry port.
     4.1.3.5  Connect the Type S pitot tube to the manometer.  Open
the Type S entry port.  Check the manometer level and zero.  Insert
and align the Type S pitot tube so that its A side impact opening is
at the same point :as was the standard pitot tube and is pointed directly
into the flow..  .Wake sure that the entry port surrounding the tube is
properly sealed.
     4.1.3.6  Read Ap  and enter its value in the data table.  Remove
the Type S pitot tube from the duct and disconnect it from the manometer.
     4.1.3.7  Repeat steps 4.1.3.3 through 4.1.3.6 above until three
pairs of Ap readings have been obtained.
     4.1.3.8  Repeat steps 4.1.3.3 through 4.1.3.7 above for the B side
of the Type S pitot tube.
     4.1.3.9  Perform calculations, as,: described in Section 4.1.4 below.
     4.1.4  Calculations.
     4.1.4.1  For each of the six pairs of Ap readings (i.e., three
from side A and three from side B) obtained in Section 4.1.3 above,
calculate the value of the Type S pitot tube coefficient as follows:

-------
                                 1-33
PITOTTUBE IDENTIFICATION NUMBER:
CALIBRATED BY: .	
DATE.

RUN NO.
1
2
3
"A" SIDE CALIBRATION
Apstd
cm H20
(in. H20)




AP(S)
cm H20
(in. H20)



Cp (SIDE A)
Cp(s)





DEVIATION
CP(S)-CP(A)





RUN NO.
1
2
3
"B"SIDE CALIBRATION
Apstd
cm H20
(in. H20)




AP(s)
cm H20
(in.H20)



Cp (SIDE B)
Cp($)





DEVIATION
Cp(s)-Cp(B)




                                  I |Cp(s)-Cp(AORB)|
   AVERAGE DEVIATION  = o(AORB) =
      MUSTBE<0.01
    Cp (SIDE A)-Cp (SIDE B) |-«-MUST BE <0.01

                 Figure 29.  Pitot tube calibration data.

-------
                                  1-34
          Cp(s) ' Cp(std)  /-^                     E«uat1on 2-
where:
     C / \   = Type S pitot tube coefficient
     C / ..N = Standard pitot tube coefficient; use 0.99 if the
               coefficient is unknown and the tube is designed
               according to the criteria of Sections 2.7.1  to 2.7.5
               of this method.
     Ap ..   = Velocity head  measured by the standard pitot tube,
               cm H20 (in.  H20)
     Ap      = Velocity head  measured by the Type  S pitot tube,
               cm H20 (in.  H20)
     4.1.4.2  Calculate C~  (side A),  the mean A-side coefficient,  and
C_ (side B), the mean B-side  coefficient; calculate the  difference
between these two average values.
     4.1.4.3  Calculate the deviation of each of the three  A-side  values
of C / i from C"  (side A),  and the deviation of each B-side value  of
C , x from £"  (side B).  Use  the following equation:
          Deviation  =  C / » -  C"  (A or B)           Equation 2-3
     4.1.4.4  Calculate o,  the average deviation from the mean, for
both the A and B sides of the pitot tube.  Use the following equation:

                       \   lCp(s)  ' S>  (AorB)|
     a (side A or B) =	=	        Eauation 2-4

-------
                               1-35
     4.1.4.5  Use the Type S pitot tube only if the values of o
(side A) and o (side B) are less than or equal  to 0.01  and if the
absolute value of the difference between C_ (A) and C"  (B) is 0.01
or less.
     4.1.5  Special considerations.
     4.1.5.1  Selection of calibration point.
     4.1.5.1.1  When an isolated Type S pitot  tube is calibrated,
select a calibration point at or near the center of the duct, and
follow the procedures outlined in Sections 4.1.3 and 4.1.4 above.
The Type S pitot coefficients so obtained, i.e., C  (side A) and
C  (side B), will be valid, so long as either:   (1) the isolated
pitot tube is used; or (2) the pitot tube is used with other com-
ponents (nozzle, thermocouple, sample probe) in an arrangement that
is free from aerodynamic interference effects  (see Figures 2-6 through
2-8).
     4.1.5.1.2  For Type S pitot tube-thermocouple combinations
(without sample probe), select a calibration point at or near the
center of the duct, and follow the procedures  outlined in Sections
4.1.3 and 4.1.4 above.  The coefficients so obtained will be valid
so long as the pitot tube-thermocouple combination is used by itself
or with other components in an interference-free arrangement (Figures
2-6 and 2-8).
     4.1.5.1.3  For assemblies with sample probes, the calibration
point should be located at or near the center of the duct; however,

-------
                          1-36
insertion of a probe sheath  into a small  duct may cause significant
cross-sectional area blockage and yield incorrect coefficient
values (Citation 9 in Section 6).  Therefore, to minimize the
blockage effect, the calibration point may be a few inches off-center
if necessary.  The actual  blockage effect will be negligible when the
theoretical blockage, as determined by a projected-area model of the
probe sheath, is 2 percent or less of the duct cross-sectional  area
for assemblies without external  sheaths (Figure 2-10a), and 3 percent
or less for assemblies with  external  sheaths (Figure 2-10b).
     4.1.5.2  For those probe assemblies in which pitot tube-nozzle
interference is a factor (i.e.,  those in which the pitot-nozzle
separation distance fails to meet the specification illustrated in
Figure 2-6a), the value of C / > depends upon the amount of free-space
between the tube and nozzle, and therefore is a function of nozzle
size.  In these instances, separate calibrations shall  be performed
with each of the commonly used nozzle sizes in place.  Note that the
single-velocity calibration technique is acceptable for this purpose,
even though the larger nozzle sizes (>0.635 cm or 1/4 in.) are not
ordinarily used for isokinetic sampling at velocities around
915 m/min (3000 ft/min), which is the calibration velocity; note
also that it is not necessary to draw an isokinetic sample during
calibration  (see Citation 19 in Section 6).
     4.1.5.3  For a probe assembly constructed such that its pitot
tube is always used in the same orientation, only one side of the
pitot tube need be calibrated (the side which will face the flow).
The pitot tube must still meet the alignment specifications of
Figure'2-2 or 2-3, however, and must have an average deviation (a)
value of 0.01 or less  (see Section 4.1.4.4).

-------
1-37
                                                       CD
                                                       in
                                                       cn
                                                       03

                                                       OJ
                                                      .0
                                                       O
                                                       +-*

                                                       '5.

                                                       "re
                                                       o

                                                       'EL
                                                      —
                                                       CD
                                                       CD

                                                       (5

                                                      •6
                                                       CD
                                                      +•>
                                                       u
                                                       CD

                                                      '
                                                      CM

                                                       £
                                                       2

-------
                              1-38
     4.1.6  Field Use and Recalibration.
     4.1.6.1  Field Use.
     4.1.6.1.1  When a Type S pitot tube  (isolated tube or assembly) is
used in the field, the appropriate coefficient value (whether assigned
or obtained by calibration) shall  be used to perform velocity calculations.
For calibrated Type S pitot tubes, the A  side coefficient shall  be used
when the A side of the tube faces  the flow, and the B side coefficient
shall be used when the B side faces the flow; alternatively, the
arithmetic average of the A and B  side coefficient values may be
used, irrespective of which side faces the flow.
     4.1.6.1.2  When a probe assembly is  used to sample a small  duct
(12 to 36 in. in diameter), the probe sheath sometimes blocks a
significant part of the duct cross-section, causing a reduction  in
the effective value of C"D(S)-  Consult Citation 9 in Section 6 for
details.  Conventional pitot-sampling probe assemblies are not recom-
mended for use in ducts having inside diameters smaller than 12  inches
(Citation 16 in Section 6).
     4.1.6.2  Recalibration.
     4.1.6.2.1  Isolated Pitot Tubes.  After each field use, the pitot
tube shall be carefully reexamined in top, side, and end views.   If the
pitot face openings are still aligned within the specifications  illus-
trated in Figure 2-2 or 2-3, it can be assumed that the baseline coef-
ficient of the pitot tube has not changed.  If, however, the tube has
been damaged to the extent that it no longer meets the specifications of
Figure 2-2 or 2-3, the damage shall either be repaired to restore
proper alignment of the face openings or the tube shall be discarded.
     4.1.6.2.2  Pitot Tube Assemblies.  After each field use, check the
face opening alignment of the pitot tube, as in Section 4.1.6.2.1; also,

-------
                              1-39
remeasure the intercomponent spacings of the assembly.  If the inter-
component spacings have not changed and the face opening alignment is
acceptable, it can be assumed that the coefficient of the assembly
has not changed.  If the face opening alignment is no longer within the
specifications of Figures 2-2 or 2-3, either repair the damage or
replace the pitot tube (calibrating the new assembly, if necessary).
If the intercomponent spacings have changed, restore the original
spacings or recalibrate the assembly.
     4.2  Standard pitot tube (if applicable).  If a standard pitot tube
is used for the velocity traverse, the tube shall be constructed
according to the criteria of Section 2.7 and shall be assigned a
baseline coefficient value of 0.99.  If the standard pitot tube is
used as part of an assembly, the tube shall be in an interference-
free arrangement (subject to the approval of the Administrator).
     4.3  Temperature Gauges.  After each field use, calibrate dial
thermometers, liquid-filled bulb thermometers, thermocouple-
potentiometer systems, and othe^ gauges at a temperature within
10 percent of the average absolute stack temperature.  For temperatures
up to 405°C (761°F), use an ASTM mercury-in-glass reference thermometer,
or equivalent, as a reference; alternatively, either a reference
thermocouple and potentiometer (calibrated by NBS) or thermometric
fixed points, e.g., ice bath and boiling water (corrected for barometric
pressure) may be used.  For temperatures above 405°C  (761°F), use an
NBS-calibrated reference thermocouple-potentiometer system or an
alternate reference, subject to the approval of the Administrator.
     If, during calibration, the absolute temperatures measured with
the gauge being calibrated and the reference gauge agree within 1.5
percent, the temperature data taken in the field shall be considered
valid.  Otherwise, the pollutant emission test shall either be

-------
                             1-40
considered invalid or adjustments (if appropriate)  of the test results
shall be made, subject to the approval  of the  Administrator.
     4.4  Barometer.   Calibrate the barometer  used  against a  mercury
barometer.
5.  Calculations
     Carry out calculations, retaining at least one extra decimal
figure beyond that of the acquired data.   Round off figures after
final calculation.
     5.1  Nomenclature.
                                            2     2
     A    = Cross-sectional  area of stack, m  (ft ).
     B    = Water vapor in the gas stream (from Method 5 or Reference
      ws
            Method 4), proportion by volume.
     C    = Pitot tube coefficient, dimensionless.
     K    = Pitot tube constant,
          ™ Q7  m   f(g/g-mo1e)(mm Hg)I
          ^'y/ sec"  |_ (*K)(mm H20)  J
                                    1/2
                         :; yum nyj i
                         i H20)  J
for the metric system and
     PR 40 -It r(lb/1b-mole)(in.  Hg)]
     00"^ sec L (°R)(1n. H20)      |
for the English system.
M.   = Molecular weight  of stack  gas,  dry basis (see Section 3.6)
       g/g-mole (Ib/lb-mole).
M    = Molecular weight  of stack  gas,  wet basis, g/g-mole
       (Ib/lb-mole).
     = Md(l - Bws) + 18.0 Bws                     Equation 2-5
Pbar = Barometric pressure at  measurement site, mm Hg (in.  Hg).
P    = Stack static pressure,  mm  Hg (in.  Hg).

-------
                              1-41
      P    = Absolute stack  gas  pressure, mm  Hg  (in.  Hg).
       d
          = Pbar * Pq                                 Equation 2-6
     Pstd = standard absolute pressure,  760 mm Hg  (29.92  in. Hg).
     Qsd  = Dry volumetric stack gas  flow rate corrected  to standard
        »
            conditions, dscm/hr  (dscf/hr).
     t$   = Stack temperature, °C (°F).
     TS   = Absolute stack temperature,  °K  (°R).
          = 273 + t  for metric                        Equation 2-7
                   o
          = 460 + t  for English                      Equation 2-8
     Tstd = standard absolute temperature,  293 °K  (528°R).
     v    = Average stack gas velocity,  m/sec (ft/sec).
     Ap   = Velocity head of stack gas,  mm  H20 (in.  FLO).
     3600 = Conversion factor, sec/hr.
     18.0 = Molecular weight of  water,  g/g-mole  (Ib/lb-mole).
     5.2  Average stack gas  velocity.
          vs  =  KpCp   (/ZF)                         Equation  2-9
     5.3  Average stack gas dry volumetric  flow rate.
     Qsd =  3600 (1-B  ) v A  L  std ) (p-S-l        Equation  2-10
      S              WS   S   Vs(aVg)y \pstd/
8.  Bibliography
     1.  Mark, L. S.   Mechanical Engineers'  Handbook.   New York.
McGraw-Hill  Book Co., Inc.  1951.
     2.  Perry, 0.  H.  Chemical Engineers'  Handbook.   New York
McGraw-Hill  Book Co., Inc.  1960.

-------
                               1-42
     3.  Shigehara, R.  T.,  W.  F.  Todd,  and W.  5.  Smith.   Significance
of Errors in Stack Sampling Measurements.   U.  S.  Environmental
Protection Agency, Research Triangle Park, N.  C.   (Presented at the
Annual Meeting of the Air  Pollution Control  Association,  St. Louis, Mo.,
June 14-19, 1970.)
     4.  Standard Method for Sampling Stacks for  Particulate Matter.
In:  1971 Book of ASTM Standards, Part  23.  Philadelphia, Pa.  1971.
ASTM Designation D-2928-71.
     5.  Vennard, J. K.  Elementary Fluid  Mechanics.   New York.
John Wiley and Sons, Inc.   1947.
     6.  Fluid Meters - Their Theory and Application.   American Society
of Mechanical Engineers, New York, N.Y.  1959.
     7.  ASHRAE Handbook of Fundamentals.   1972.   p.  208.
     8.  Annual Book of ASTM Standards, Part 26.   1974.   p.  648.
     9.  Vollaro, R. F.  Guidelines for Type S Pitot  Tube Calibration.
U. S. Environmental Protection Agency,  Research Triangle  Park,  N. C.
(Presented at 1st Annual Meeting, Source Evaluation Society, Dayton, Ohio,
September 18, 1975.)
     10.  Vollaro, R. F.  A Type S Pitot Tube Calibration Study.  U. S.
Environmental Protection Agency, Emission Measurement Branch, Research
Triangle Park, N. C.  July 1974.
     11.  Vollaro, R. F.  The Effects of Impact Opening Misalignment on
the Value of the Type S Pitot Tube Coefficient. U. S. Environmental
Protection Agency, Emission Measurement Branch, Research Triangle Park,
N. C.  October 1976.

-------
                               1-43
     12.  Vollaro, R.  F.   Establishment of a Baseline Coefficient Value
for Properly Constructed  Type S Pi tot Tubes.  U.  S.  Environmental
Protection Agency, Emission Measurement Branch, Research Triangle
Park, N. C.  November  1976.
     13.  Vollaro, R.  F.   An Evaluation of Single-Velocity Calibration
Technique as a Means of Determining Type S Pitot  Tube Coefficients.
U. S. Environmental Protection Agency, Emission Measurement Branch,
Research Triangle Park, N. C.  August 1975.
     14.  Vollaro, R.  F.   The Use of Type S Pitot Tubes for the  Measurement
of Low Velocities.  U. S. Environmental Protection Agency, Emission
Measurement Branch, Research Triangle Park, N.  C.  November 1976.
     15.  Smith, Marvin L.  Velocity Calibration  of EPA Type Source
Sampling Probe.  United Technologies Corporation, Pratt and Whitney
Aircraft Division, East Hartford, Conn.  1975.
     16.  Vollaro, R.  F.   Recommended Procedure for Sample Traverses  in
Ducts Smaller than 12  Inches in Diameter.  U. S.  Environmental  Protection
Agency, Emission Measurement Branch, Research Triangle Park, N.  C.
November 1976.
     17.  Ower, E. and R. C. Pankhurst.  The Measurement of Air  Flow,
4th Ed. London, Pergamon  Press.  1966.
     18.  Vollaro, R.  F.   A Survey of Commercially Available Instrumentation
for the Measurement of Low-Range Gas Velocities.   U. S. Environmental
Protection Agency, Emission Measurement Branch, Research Triangle Park,
N. C.  November 1976.   (Unpublished Paper)

-------
                                1-44
     19.  Gnyp, A.  W.,  C.  C.  St.  Pierre,  D.  S.  Smith,  D.  Mozzon,  and
J. Steiner.   An Experimental  Investigation of the Effect  of Pi tot Tube-
Sampling Probe Configurations on  the Magnitude of the  S Type Pitot Tube
Coefficient for Commercially Available Source Sampling Probes.   Prepared
by the University of Windsor for  the Ministry of the Environment,
Toronto, Canada.  February 1975.

-------
GAS ANALYSIS FOR CARBON DIOXIDE, EXCESS AIR,



       AND DRY MOLECULAR WEIGHT





               METHODS

-------
                              1-45
        METHOD 3--GAS ANALYSIS FOR CARBON DIOXIDE,  OXYGEN,
              EXCESS AIR, AND DRY MOLECULAR WEIGHT
 1.   Principle and Applicability
     1.1  Principle.   A gas sample is extracted  from a stack,  by one
of the following methods:   (1) single-point, grab sampling;  (2)  single-
point, integrated sampling; or (3) multi-point,  integrated sampling.
The gas sample is analyzed for percent carbon dioxide (C02),  percent
oxygen (02), and, if necessary, percent carbon monoxide (CO).   If a
dry molecular weight determination is to  be made, either an Orsat
or a Fyrite  analyzer may be used for the analysis;  for excess air
or emission rate correction factor determination, an Orsat analyzer
must be used.
     1.2  Applicability.  This method is  applicable  for determining
C02 and 02 concentrations, excess air, and dry molecular weight  of
a sample from a gas stream of a fossil-fuel combustion process.   The
method may also be applicable to other processes where it has  been
determined that compounds other than CO^, 02> CO, and nitrogen (N2)
are not present in concentrations sufficient to  affect the results.
     Other methods, as well as modifications to  the  procedure
described herein, are also applicable for some or all of the  above
determinations.  Examples of specific methods and modifications
include:  (1) a multi-point sampling method using an Orsat analyzer
to analyze individual grab samples obtained at each  point; (2) a
method using C02 or 02 and stoichiometric calculations to determine
dry molecular weight and excess air; (3)  assigning a value of 30.0
 Mention of trade names or specific products does not constitute
 endorsement by the Environmental Protection Agency.

-------
                                1-46
 for dry molecular weight,  in  lieu of actual measurements, for
 processes  burning natural  gas, coal, or oil.  These methods and
 modifications  may be  used, but are  subject to the  approval of  the
 Administrator, U.S. Environmental Protection Agency.
 2.   Apparatus
      As an alternative  to  the sampling apparatus and  systems described
 herein, other  sampling  systems (e.g., liquid displacement) may be
 used provided  such  systems are capable of obtaining a representative
 sample  and maintaining  a constant sampling rate, and  are otherwise
 capable of yielding acceptable results.  Use of such  systems is
 subject to the approval of the Administrator.
      2.1  Grab Sampling (Figure  3-1).
      2.1.1  Probe.  The probe should be made of stainless steel  or
 borosilicate glass  tubing  and should be equipped with an in-stack
 or out-stack filter to  remove particulate matter (a plug of glass
*wool is satisfactory  for this purpose).  Any other material inert to
 CL, CCL, CO, and N« and resistant to temperature at sampling conditions
 may be  used for the probe; examples of such material  are aluminum,
 copper, quartz glass  and Teflon.
      2.1.2  Pump.   A  one-way  squeeze bulb, or equivalent, is used
 to transport the gas  sample to the  analyzer.
      2.2  Integrated  Sampling (Figure 3-2).
      2.2.1  Probe.  A probe such as that described in Section  2.1.1
 is suitable.

-------
                                        1-47
                        PROBE
                                                FLEXIBLE TUBING
                FILTER (GLASS WOOL)
                                    SQUEEZE BULB
                                                                       TO ANALYZER
                                 Figure 3-1.  Grab sampling train.
                                                RATE METER
          AIR-COOLED
          CONDENSER
PROBE
    \
        FILTER
     (GLASS WOOL)
                                               QUICK DISCONNECT

                                                          J-l
                                      RIGID CONTAINER
                          Figure 3-2. Integrated gas-sampling train.

-------
                              1-48
     2.2.2  Condenser.   An air-cooled or water-cooled  condenser,
or other condenser that will  not remove 02,  C02,  CO, and  N2,  may
be used to remove excess moisture which would  interfere with
the operation of the pump and flow meter.
     2.2.3  Valve.  A needle  valve is used to  adjust sample gas
flow rate.
     2.2.4  Pump.  A leak-free,  diaphragm-type pump, or equivalent,
is used to transport sample gas  to the flexible bag.   Install  a
small surge tank between the  pump and rate meter  to eliminate the
pulsation effect of the diaphragm pump on the  rotameter.
     2.2.5  Rate Meter.  The  rotameter, or equivalent  rate meter,
used should be capable  of measuring flow rate  to  within +2 percent
of the selected flow rate. A flow rate range  of  500 to 1000
cm /nrin is suggested.
     2.2.6  Flexible Bag.  Any leak-free plastic  (e.g., Tedlar,
Mylar, Teflon) or plastic-coated aluminum (e.g.,  aluminized
Mylar) bag, or equivalent, having a capacity consistent with  the
selected flow rate and time length of the test run, may be used.  A
capacity in the range of 55 to 90 liters is  suggested.
     To leak-check the bag, connect it to a  water manometer and  pressurize
the bag to 5 to 10 cm H20 (2  to 4 in. HJ)).  Allow to  stand for  10
minutes.  Any displacement in the water manometer indicates a
leak.  An alternative leak-check method is  to  pressurize  the  bag
to 5 to 10 cm H20 (2 to 4 in. H20) and allow to stand  overnight.   A
deflated bag indicates a leak.

-------
                             1-49
     2.2.7  Pressure Gauge.   A water-filled U-tube manometer,  or
equivalent, of about 28 cm (12 in.) is used for the flexible bag
leak-check.
     2.2.8  Vacuum Gauge.   A mercury manometer, or equivalent, of at
least 760 mm Hg (30 in. Hg)  is used for the sampling train leak-check.
     2.3  Analysis.  For Orsat and Fyrite analyzer maintenance and
operation procedures, follow the instructions recommended by the
manufacturer, unless otherwise specified herein.
     2.3.1  Dry Molecular Weight Determination.  An Orsat analyzer or
Fyrite type combustion gas analyzer may be used.
     2.3.2  Emission Rate Correction Factor or Excess Air Determination.
An Orsat analyzer must be used.  For low (XL (less than 4.0 percent)
or high Op (greater than 15.0 percent) concentrations, the measuring
burette of the Orsat must have at least 0.1 percent subdivisions.
3.  Dry Molecular Weight Determination
     Any of the three sampling and analytical procedures described
below may be used for determining the dry molecular weight.
     3.1  Single-Point, Grab Sampling and Analytical Procedure.
     3.1.1  The sampling point in the duct shall either be at  the
centroid of the cross section or at a point no closer to the walls
than 1.00 m (3.3 ft), unless otherwise specified by the Administrator.
     3.1.2  Set up the equipment as shown in Figure 3-1, making sure
all connections ahead of the analyzer are tight and leak-free.  If an
Orsat analyzer is used, it is recommended that the analyzer be leak-
checked by following the procedure in Section 5; however, the  leak-
check is optional.

-------
                             1-50
     3.1.3  Place the probe in the stack,  with  the  tip  of the  probe
positioned at the sampling point;  purge the  sampling  line.   Draw a
sample into the analyzer and immediately analyze  it for percent C0?
and percent 0_.  Determine the percentage  of the  gas  that is N2 and
CO by subtracting the sum of the percent CO. and  percent 0?  from
100 pet  ;nt.   Calculate the dry molecular  weight  as indicated  in
Section 6.3.
     3.1.4  Repeat the sampling, analysis, and  calculation procedures,
until the dry molecular weights of any three grab samples differ from
their mean by no more than 0.3 g/g-mole (0.3 Ib/lb-mole).  Average
these three molecular weights, and report  the results to the nearest
0.1 g/g-mole (Ib/lb-mole).
     3.2  Single-Point, Integrated Sampling  and Analytical Procedure.
     3.2.1  The sampling point in the duct shall  be located  as speci-
fied in Section 3.1.1.
     3.2.2  Leak-check (optional) the flexible  bag  as in Section 2.2.6.
Set up the equipment as shown in Figure 3-2. Just  prior to  sampling,
leak-check (optional) the train by placing a vacuum gauge at the
condenser inlet, pulling a vacuum of at least 250 mm  Hg (10  in. Hg),
plugging the outlet at the quick disconnect, and  then turning  off the
pump.  The vacuum should remain stable for at least 0.5 minute.
Evacuate the flexible bag.  Connect the probe and place it in  the
stack, with the tip of the probe positioned  at  the  sampling  point;
purge the sampling line.  Next, connect the  bag and make sure  that
all connections are tight and leak free.

-------
                              1-51
     3.2.3  Sample at a constant rate.  The sampling run should be
simultaneous with, and for the same total  length of time as, the
pollutant emission rate determination.  Collection of at least 30
               o
liters (1.00 ft ) of sample gas is recommended; however, smaller
volumes may be collected, if desired.
     3.2.4  Obtain one integrated flue gas sample during each pollutant
emission rate determination.  Within 8 hours after the sample is taken,
analyze it for percent COp and percent 0,, using either an Orsat
analyzer or a Fyrite type combustion gas analyzer.  If an Orsat
analyzer is used, it is recommended that the Orsat leak-check
described in Section 5 be performed before this determination;
however, the check is optional.  Determine the percentage of the
gas that is Np and CO by subtracting the sum of the percent COp and
percent 02 from 100 percent.  Calculate the dry molecular weight as
indicated in Section 6.3.
     3.2.5  Repeat the analysis and calculation procedures until the
individual dry molecular weights for any three analyses differ from
their mean by no more than 0.3 g/g-mole (0.3 Ib/lb-mole).  Average
these three molecular weights, and report the results to the nearest
0.1 g/g-mole (0.1 Ib/lb-mole).
     3.3. Multi-Point, Integrated Sampling and Analytical Procedure.
     3.3.1  Unless otherwise specified by the Administrator, EPA, a
minimum of eight traverse points shall be used for circular stacks having
diameters less than 0.61 m (24 in.), a minimum of nine shall be used for
rectangular stacks having equivalent diameters less than 0.61 m (24 in.),
and a minimum of twelve traverse points shall be used for all other
cases.  The traverse points shall be located according to Method 1.  The
use of fewer points is subject to the approval of the Administrator.

-------
                            1-52
     3.3.2  Follow the procedures outlined in Sections 3.2.2
through 3.2.5, except for the following:   traverse all sampling
points and sample at each point for an equal  length of time.
Record sampling data as shown in Figure 3-3.
4.  Emission Rate Correction Factor or Excess Air Determination
     Note:  A Fyrite type combustion gas  analyzer is not acceptable
 or excess air or emission rate correction factor determination,
unless approved by the Administrator,   If both percent C02 and
percent CL are measured, the analytical results of any of the three
procedures given below may also be used for calculating the dry
molecular weight.
     Each of the three procedures below shall be used only when
specified in an applicable subpart of the standards.  The use of
these procedures for other purposes must  have specific prior approval
of the Administrator.
     4.1  Single-Point, Grab Sampling and Analytical Procedure.
     4.1.1  The sampling point in the duct shall either be at the
centroid of the cross-section or at a point no closer to the walls
than 1.00 m (3.3 ft), unless otherwise specified by the Administrator.
     4.1.2  Set up the equipment as shown in Figure 3-1, making sure
all connections ahead of the analyzer are tight and leak-free.  Leak-
check the Orsat analyzer according to the procedure described in
Section 5.  This leak-check is mandatory.
     4.1.3  Place the probe in the stack, with the tip of the probe
positioned at the sampling point; purge the sampling line.  Draw

-------
                 1-53
TIME




TRAVERSE
PT.




AVERAGE
Q
1pm





% DEV.a





a
%DEV=
                         (MUST BE < 10%)
     Figure 3-3. Sampling rate data.

-------
                              1-54
a sample into the analyzer.   For emission rate correction factor
determination, immediately analyze the sample, as  outlined in  Sections
4.1.4 and 4.1.5, for percent C0? or percent 02.   If excess air is
desired, proceed as follows:  (1) immediately analyze the sample,  as in
Sections 4.1.4 and 4.1.5, for percent CO,,, 02, and CO; (2) determine the
percentage of the gas that is Np by subtracting  the sum of the percent
C02, percent 0,,, and percent CO from 100 percent;  and (3) calculate
percent excess air as outlined in Section 6.2.
     4.1.4  To ensure complete absorption of the C02, 0^, or if appli-
cable, CO, make repeated passes through each absorbing solution until
two consecutive readings are the same.  Several  passes (three  or four)
should be made between readings.  (If constant readings cannot be
obtained after three consecutive readings, replace the absorbing solution.)
     4.1.5  After the analysis is completed, leak-check (mandatory)
the Orsat analyzer once again, as described in Section 5.  For the
results of the analysis to be valid, the Orsat analyzer must pass
this leak test before and after the analysis.  Note:  Since this
single-point, grab sampling and analytical procedure is normally
conducted in conjunction with a single-point, grab sampling and
analytical procedure for a pollutant, only one analysis is ordinarily
conducted.  Therefore, great care must be taken to obtain a valid
sample and analysis.  Although in most cases  only C02 or 0? is
required, it is recommended that both C02 and 02 be measured,  and
that Citation 5 in the Bibliography be used to validate the analytical
data.

-------
                             1-55

     4.2  Single-Point, Integrated  Sampling  and  Analytical  Procedure.

      .2.1   The sampling point in the duct shall  be located as

specified in Section 4.1.1.

     4.2.2   Leak-check (mandatory)  the flexible  bag as  in  Section  2.2.6.

Set up the  equipment as shown in Figure 3-2.   Just prior  to sampling,

leak-check  (mandatory) the train by placing  a  vacuum gauge  at the

condenser inlet, pulling a vacuum of at least  250 mm Hg (10 in.  Hg),

plugging the outlet at the quick disconnect, and then turning off  the

pump.  The  vacuum shall remain stable for at least 0.5  minute.

Evacuate the flexible bag.  Connect the probe  and place it  in the

stack, with the tip of the probe positioned  at the sampling point;

purge the sampling line.  Next, connect the  bag  and make  sure that all

connections are tight and leak free.

     4.2.3   Sample at a constant rate, or as specified  by  the

Administrator.   The sampling run must be simultaneous with, and  for

the same total  length of time as, the pollutant  emission  rate
                                                   o
determination.   Collect at least 30 liters (1.00 ft ) of  sample

gas.  Smaller volumes may be collected, subject  to approval of the

Administrator.

     4.2.4  Obtain one integrated flue gas sample during  each pollutant

emission rate determination.  For emission rate  correction  factor

determination,  analyze the sample within 4 hours after it  is taken

for percent CO^ or percent 0^ (as outlined in  Sections  4.2.5 through

4.2.7).  The Orsat analyzer must be leak-checked (see Section 5) before

the analysis.  If excess air is desired, proceed as follows:  (1)  within

4 hours after the sample is taken,  analyze it  (as in Sections 4.2.5

-------
                                1-56
through 4.2.7) for percent C02, 0^, and CO;  (2) determine the percentage
of the gas that is ^ by subtracting the sum of the percent CO.,, percent
Op, and percent CO from 100 percent; (3) calculate percent excess air,
as outlined in Section 6.2.
     4.2.5  To ensure complete absorption of the COp, Op, or if applicable,
CO, make repeated passes through each absorbing solution until two
consecutive readings are the same.  Several passes (three or four)
should be made between readings.  (If constant readings cannot be obtained
after three consecutive readings, replace the absorbing solution.)
     4.2.6  Repeat the analysis until the following criteria are met:
     4.2.6.1  For percent COp, repeat the analytical procedure until the
results of any three analyses differ by no more than (a) 0.3 percent by
volume when COp is greater than 4.0 percent or (b) 0.2 percent by volume
when COp is less than or equal to 4.0 percent.   Average the three acceptable
values of percent COp and report the results tu the nearest 0.1  percent.
     4.2.6.2  For percent Op, repeat the analytical procedure until  the
results of any three analyses differ by no more than (a) 0.3 percent by
volume when D£ is less than 15.0 percent or (b) 0.2 percent by volume
when Op is greater than or equal to 15.0 percent.   Average the three
acceptable values of percent Op and report the results to the nearest
0.1 percent.
     4.2.6.3  For percent CO, repeat the analytical procedure until  the
results of any three analyses differ by no more than 0.3 percent.
Average the three acceptable values of percent CC  and report the results
to the nearest 0.1 percent.

-------
                              1-57
     4.2.7  After the analysis is completed, leak-check (mandatory)
the Orsat analyzer once again, as described in Section 5.   For the
results of the analysis to be valid, the Orsat analyzer must pass
this leak test before and after the analysis.  Note:  Although in
most instances only CCL or 0. is required, it is recommended that
both C0? and CL be measured, and that Citation 5 in the Bibliography
be used to validate the analytical data.
     4.3  Multi-Point, Integrated Sampling and Analytical  Procedure.
     4.3.1  Both the minimum number of sampling points and the
sampling point location shall be as specified in Section 3.3.1 of
this method.  The use of fewer points than specified is subject to
the approval of the Administrator.
     4.3.2  Follow the procedures outlined in Sections 4.2.2 through
4.2.7, except for the following:  Traverse all sampling points and
sample at each point for an equal length of time.  Record  sampling
data as shown in Figure 3-3.
5.  Leak-Check Procedure for Orsat Analyzers
     Moving an Orsat analyzer frequently causes it to leak.  Therefore,
an Orsat analyzer should be thoroughly leak-checked on site before the
flue gas sample is introduced into it.  The procedure for leak-checking
an Orsat analyzer is:
     5.1.1  Bring the liquid level in each pipette up to the reference
mark on the capillary tubing and then close the pipette stopcock.
     5.1.2  Raise the leveling bulb sufficiently to bring the confining
liquid meniscus onto the graduated portion of the burette and then
close the manifold stopcock.

-------
                              1-58



     5.1.3  Record the meniscus position.



     5.1.4  Observe the meniscus in  the burette and the liquid level



in the pipette for movement over the next  4 minutes.



     5.1.5  For the Orsat analyzer to pass the leak-check, two



conditions must be met:



     5.1.5.1   The liquid level  in each pipette must not fall  below



the bottom of the capillary tubing during  this 4-mini/te interval.



     5.1.5.2  The meniscus in the burette  must not change by more



than 0.2 ml during this 4-minute interval.



     5.1.6  If the analyzer fails the leak-check procedure, all



rubber connections and stopcocks should be checked until the cause



of the leak is identified.  Leaking stopcocks must be disassembled,



cleaned, and regreased.  Leaking rubber connections must be replaced.



After the analyzer is reassembled, the leak-check procedure must be



repeated.



6.  Calculations



     6.1  Nomenclature.



     M,    = Dry molecular weight, g/g-mole (Ib/lb-mole).



     %EA   = Percent excess air.



     %COp  = Percent CO^ by volume (dry basis).



     %0?   = Percent 00 by volume (dry basis).



     %CO   = Percent CO by volume (dry basis).



     %N?   = Percent N? by volume (dry basis).



     0.264 = Ratio of 02 to N2  in air, v/v.

-------
                                1-59
     0.280 = Molecular weight of N2 or CO, divided by 100.
     0.320 = Molecular weight of Op divided by 100.
     0.440 = Molecular weight of COp divided by 100.
     6.2  Percent Excess Air.  Calculate the percent excess air (if
applicable), by substituting the appropriate values of percent Op,
CO, and N~ (obtained from Section 4.1.3 or 4.2.4) into Equation 3-1.
     %EA  =
                        - 0.5% CO
100       Equation 3-1
              0.264 %N2 - (%02 - 0.5% CO)
              _»                          .MH
     Note:  The equation above assumes that ambient air is used as
the source of Op and that the fuel does not contain appreciable
amounts of Np (as do coke oven or blast furnace gases).   For those
cases when appreciable amounts of Np are present (coal,  oil, and
natural gas do not contain appreciable amounts of Np)  or when oxygen
enrichment is used, alternate methods, subject to approval of the
Administrator, are required.
     6.3  Dry Molecular Weight.   Use Equation 3-2 to calculate the
dry molecular weight of the stack gas.

     Md  =  0.440(%C02) + 0.320(%02) + 0.280(%N2 + %CO)    Equation  3-2
     Note:  The above equation does not consider argon in air (about
0.9 percent, molecular weight of 37.7).   A negative error of about
0.4 percent is introduced.   The  tester may opt to include argon in
the analysis using procedures subject to approval of the Administrator.

-------
                              1-60
7.  Bibliography
     1.  Altshuller, A.P.   Storage of Gases and Vapors in Plastic
Bags.  International Journal  of Air and Water Pollution.   6_:75-81.
1963.
     2.  Conner, William D.  and J. S. Nader.   Air Sampling with Plastic
Bags.  Journal  of the American Industrial  Hygiene Association.
25_:291-297.  1964.
     3.  Burrell Manual  for  Gas Analysts,  Seventh edition.  Burrell
Corporation, 2223 Fifth  Avenue, Pittsburgh, Pa. 15219.  1951.
     4.  Mitchell,  W.J.  and  M.R. Midgett.   Field Reliability of the
Orsat Analyzer.  Journal of  Air Pollution  Control Association.
26:491-495.  May 1976.
     5.  Shigehara, R.T.,  R.M. Neulicht, and W.S. Smith.   Validating
Orsat Analysis  Data from Fossil Fuel-Fired Units.  Stack  Sampling
News.  4(2):21-26.   August,  1976.

-------
DETERMINATION OF MOISTURE IN STACK GASES





                METHOD 4

-------
                             1-61
           METHOD A—DETERMINATION OF MOISTURE CONTENT
                        IN STACK GASES
1.   Principle and Applicability
     1.1  Principle.  A gas sample is extracted at a constant rate from
the source; moisture is removed from the sample stream and determined
either volumetrically or gravimetrically.
     1.2  Applicability.  This method is applicable for determining the
moisture content of stack gas.
     Two procedures are given.  The first is a reference method,  for
accurate determinations of moisture content (such as are needed to
calculate emission data).  The second is an approximation method, which
provides estimates of percent moisture to aid in setting isokinetic
sampling rates prior to a pollutant emission measurement run.  The
approximation method described herein is only a suggested approach;
alternative means for approximating the moisture content, e.g., drying
tubes, wet bulb-dry bulb techniques, condensation techniques, stoichio-
metric calculations, previous experience, etc., are also acceptable.
     The reference method is often conducted simultaneously with
a pollutant emission measurement run; when it is, calculation of percent
isokinetic, pollutant emission rate, etc., for the run shall be based upon
the results of the reference method or its equivalent; these calculations
shall not be based upon the results of the approximation method,  unless
the approximation method is shown, to the satisfaction of the Administrator,
U.  S. Environmental Protection Agency, to be capable of yielding results
within 1 percent HpO of the reference method.

-------
                               1-62
     Note:  The reference method  may yield questionable results
when applied to saturated gas streams or to streams that contain
water droplets.  Therefore, when  these conditions exist or are
suspected, a second determination of the moisture content shall
be made simultaneously with the reference method, as follows:   Assume
that the gas stream is saturated.  Attach a temperature sensor [capable
of measuring to + 1° C (2° F)]  to the reference method probe.   Measure
the stack gas temperature at each traverse point (see Section  2.2.1)
during the reference method traverse; calculate the average stack gas
temperature.  Next, determine the moisture percentage, either  by:
(1) using a psychrometric chart and making appropriate corrections if
stack pressure is different from  that of the chart, or (2) using
saturation vapor pressure tables.  In cases where the psychrometric
chart or the saturation vapor pressure tables are not applicable (based
on evaluation of the process),  alternate methods, subject to the approval
of the Administrator, shall be  used.
2.  Reference Method
     The procedure described in Method 5 for determining moisture
content is acceptable as a reference method.
     2.1  Apparatus.  A schematic of the sampling train used in
this reference method is shown  in Figure 4-1.  All  components
shall be maintained and calibrated according to the procedure
outlined in Method 5.
     2.1.1  Probe.  The probe is  constructed of stainless steel  or
glass tubing, sufficiently heated to prevent water  condensation,  and
is equipped with a filter, either in-stack (e.g., a plug of glass

-------
1-63
                                         O
                                         QJ
                                         u
                                         E
                                         «
                                         (A

                                         0)
                                         o
                                         2


-------
                              1-64
wool inserted into the end of the probe)  or heated  out-stack (e.g.,
as described in Method 5), to remove participate matter.
     When stack conditions permit, other  metals or  plastic tubing
may be used for the probe, subject to the approval  of the
Administrator.
     2.1.2  Condenser.  The condenser consists of four impingers
connected in series with ground glass, leak-free fittings or any
similarly leak-free non-contaminating fittings.  The first, third,  and
fourth impingers shall be of the Greenburg-Smith design,  modified
by replacing the tip with a 1.3 centimeter (1/2 inch) ID  glass tube
extending to about 1.3 cm (1/2 in.)  from  the bottom of the flask.
The second impinger shall be of the Greenburg-Smith design with the
standard tip.  Modifications (e.g., using flexible  connections
between the impingers, using materials other than glass,  or using
flexible vacuum lines to connect the filter holder  to the condenser)
may be used, subject to the approval of the Administrator.
     The first two impingers shall contain known volumes  of water,
the third shall be empty, and the fourth  shall contain a  known
weight of 6- to 16-mesh indicating type silica gel, or equivalent
desiccant.  If the silica gel has been previously used, dry at
175°C (350°F) for 2 hours.  New silica gel may be used as received.
A thermometer, capable of measuring temperature to  within 1°C
(2°F), shall be placed at the outlet of the fourth  impinger, for
monitoring purposes.
      Alternatively,  any  system may  be used  (subject  to the approval
of  the Administrator) that  cools  the  sample gas stream and allows

-------
                                1-65
measurement of both the water that has been condensed and the
moisture leaving the condenser, each to within 1  ml or 1 g.
Acceptable means are to measure the condensed water, either gravi-
metrically or volumetrically, and to measure the  moisture leaving
the condenser by:  (1) monitoring the temperature and pressure at
the exit of the condenser and using Dalton's law  of partial pressures,
or (2) passing the sample gas stream through a tared silica gel  (or
equivalent desiccant) trap, with exit gases kept  below 20°C (68°F), and
determining the weight gain.
     If means other than silica gel are used to determine the amount of
moisture leaving the condenser, it is recommended that silica gel (or
equivalent) still be used between the condenser system and pump, to
prevent moisture condensation in the pump and metering devices and to
avoid the need to make corrections for moisture in the metered volume.
     2.1.3  Cooling System.  An ice bath container and crushed ice
(or equivalent) are used to aid in condensing moisture.
     2.1.4  Metering System.  This system includes a vacuum gauge,
leak-free pump, thermometers capable of measuring temperature to
within 3°C (5.4°F), dry gas meter capable of measuring volume to
within 2 percent, and related equipment as shown  in Figure 4-1.
Other metering systems, capable of maintaining a  constant sampling
rate and determining sample gas volume, may be used, subject to the
approval of the Administrator.
     2.1.5  Barometer.  Mercury, aneroid, or other barometer
capable of measuring atmospheric pressure to within 2.5 mm Hg (0.1
in. Hg) may be used.  In many cases, the barometric reading may be
obtained from a nearby national weather service station, in which

-------
                               1-66
case the station value (which is the absolute barometric pressure)
shall be requested and an adjustment for elevation differences
between the weather station and the sampling point shall be applied
at a rate of minus 2.5 mm Hg (0.1 in. Hg) per 30 m (100 ft) eleva-
tion increase or vice versa for elevation decrease.
     2.1.6  Graduated Cylinder and/or Balance.  These items are
used to measure condensed water and moisture caught in the silica
gel to within 1 ml or 0.5 g.  Graduated cylinders shall have
subdivisions no greater than 2 ml.  Most laboratory balances are
capable of weighing to the nearest 0.5 g or less.  These balances
are suitable for use here.
     2.2  Procedure.  The following procedure is written for a
condenser system (such as the impinger system described in Section
2.1.2) incorporating volumetric analysis to measure the condensed
moisture, and silica gel and gravimetric analysis to measure the
moisture leaving the condenser.
     2.2.1  Unless otherwise specified by the Administrator, a minimum of
eight traverse points shall be used for circular stacks having diameters
less than 0.61 m (24 in.), a minimum of nine points shall be used for
rectangular stacks having equivalent diameters less than 0.61 m (24 in.),
and a minimum of twelve traverse points shall be used in all other cases.
The traverse points shall be located according to Method 1.  The use of
fewer points is subject to the approval of the Administrator.  Select a
suitable probe and probe length such that all traverse points can be
sampled.  Consider sampling from opposite sides of the stack (four total
sampling ports) for large stacks, to permit use of shorter probe lengths.
Mark the probe with heat resistant tape or by some other method to

-------
                              1-67
denote the proper distance into  the stack or duct for each sampling
point.  Place known volumes of water in the first two impingers.   Weigh
and record the weight of the silica gel to the nearest 0.5 g, and
transfer the silica gel to the fourth impinger; alternatively, the
silica gel may first be transferred to the impinger, and the weight of
the silica gel plus impinger recorded.
     2.2.2  Select a total sampling time such that a minimum total gas
volume of 0.60 scm (21 scf) will be collected, at a rate no greater than
0.021 m /min (0.75 cfm).  When both moisture content and pollutant
emission rate are to be determined, the moisture determination shall be
simultaneous with, and for the same total length of time as, the pollutant
emission rate run, unless otherwise specified in an applicable subpart of
the standards.
     2.2.3  Set up the sampling  train as shown in Figure 4-1.  Turn on
the probe heater and (if applicable) the filter heating system to
temperatures of about 120°C (248°F), to prevent water condensation
ahead of the condenser; allow time for the temperatures to stabilize.
Place crushed ice in the ice bath container.  It is recommended, but
not required, that a leak check be done, as follows:  Disconnect the
probe from the first impinger or (if applicable) from the filter holder.
Plug the inlet to the first impinger (or filter holder) and pull  a
380 mm (15 in.) Hg vacuum; a lower vacuum may be used, provided that
it is not exceeded during the test.  A leakage rate in excess of
                                                   o
4 percent of the average sampling rate or 0.00057 m /min (0.02 cfm),
whichever is less, is unacceptable.  Following the leak check,
reconnect the probe to the sampling train.

-------
                               1-68
     2.2.4  During the sampling run, maintain a sampling rate
within 10 percent of constant rate, or as specified by the
Administrator.   For each run, record the data required on the
example data sheet shown in Figure 4-2.   Be sure to record the
dry gas meter reading at the beginning and end of each sampling
time increment and whenever sampling is  halted.  Take other
appropriate readings at each sample point, at least once during
each time increment.
     2.2.5  To begin sampling, position  the probe tip at the first
traverse point.  Immediately start the pump and adjust the flow to the
desired rate.  Traverse the cross section, sampling at each traverse
point for an equal length of time.  Add  more ice and, if necessary, salt
to maintain a temperature of less than 20°C (68°F) at the silica gel
outlet.
     2.2.6  After collecting the sample, disconnect the probe from the
filter holder (or from the first impinger) and conduct a leak check
(mandatory) as described in Section 2.2.3.  Record the leak rate.  If
the leakage rate exceeds the allowable rate, the tester shall either
reject the test results or shall correct the sample volume as in
Section 6.3 of Method 5.  Next, measure the volume of the moisture
condensed to the nearest ml.  Determine the increase in weight of the
silica gel (or silica gel plus impinger) to the nearest 0.5 g.  Record
this information  (see example data sheet, Figure 4-3) and calculate the
moisture percentage, as described in 2.3 below.
     2.3  Calculations.  Carry out the following calculations, retaining
at  least one extra decimal figure beyond that of the acquired data.
Round off figures after final calculation.

-------
                  1-69
     £ ^

     Hi
&Siu.*&i£
5 s 5 s 5
i 2 & S S


















o
K
UJ
M
0
K
U
if
u
K
Cfl
u.
e
u
<
UJ
jj








UJ OC CC
ec o uj
itfiie
UJ U. < U
J S u
H- (J _l
UJ
OC
Is
flc jr
Ul
M
»" O
l!
M <
s
(3








a
uj £
5
!•» yj
u.
1 1»
= -£
0 g
e
^" ej
uJe

~ i"
K


n"
E £
< ""e


i»"

t j — • *
" o*^
o

«J C

E
0>
u
r-
§
0)
O)
k.
c
g
c
|
QJ
S
T3
01
w
3
*-•
in
'6
E
T5
O>
LL

-------
                  1-70

FINAL
INITIAL
DIFFERENCE
IMPINGER
VOLUME,
ml



SILICA GEL
WEIGHT.
9



Figure 43.  Analytical data - reference method.

-------
                         1-71
2.3.1  Nomenclature.
B     = Proportion of water vapor, by volume, in the gas stream.
MW    = Molecular weight of water, 18.0 g/g-mole (18.0 Ib/lb-mole).
P     = Absolute pressure (for this method, same as  barometric
        pressure) at the dry gas meter, mm Hg (in.  Hg).
Pstd  = standard absolute pressure, 760 mm Hg (29.92 in. Hg).
R     = Ideal gas constant, 0.06236 (mm Hg)(m3)/(g-mole)(°K)
        for metric units and 21.85 (in. Hg)(ft3)/(lb-mole)(°R)
        for English units.
T     = Absolute temperature at meter, °K (°R).
T td  = Standard absolute temperature, 293°K (528°R).
V     = Dry gas volume measured by dry gas meter, dcm (dcf).
AV    = Incremental dry gas volume measured by dry gas meter  at
        each traverse point, dcm (dcf).
V /  ..* = Dry gas volume measured by the dry gas meter,  corrected
        to standard conditions, dscm (dscf).
V  , ..% = Volume of water vapor condensed corrected to  standard
        conditions, scm (scf).
V    / . .\ = Volume of water vapor collected in silica  gel  corrected
        to standard conditions, scm (scf).
Vf    = Final volume of condenser water, ml.
V.    = Initial volume, if any, of condenser water,  ml.
W,.    = Final weight of silica gel or silica gel plus  impinger, g.
W.    = Initial weight of silica gel or silica gel  plus  impinger, g.
Y     = Dry gas meter calibration factor.
PW    = Density of water, 0.9982 g/ml  (0.002201  Ib/ml).

-------
                              1-72
     2.3.2  Volume of water vapor condensed.
     v        .  (WpwRTstd
     V(std) -
                     f
     where:
                   (V  - V)                         Equation 4-1
             = 0.001333 m /ml  for metric units
             = 0.04707 ft3/ml  for English units
     2.3.3  Volume of water vapor collected in silica gel.
     V
_  ("f Hl>  RTstd
      wsg(std)
                =  K2 (Wf - \i.)                  Equation  4-2
     where:
                         3
          K« = 0.001335 m /g for metric units
             = 0.04715 ft3/g for English units
     2.3.4  Sample gas volume.
                   (fV
     vm(std)= y ™—
             s K3Y  »•!!>                        Equation  4-3
                      m
     where:
          K3 = 0.3858 °K/mm Hg for metric  units
             » 17.64 °R/in. Hg for English units
     Note:  If the post-test leak rate  (Section  2.2.6) exceeds  the
allowable rate, correct the value of Vm in Equation  4-3,  as described
in Section 6.3 of Method 5.

-------
                                1-73
     2.3.5  Moisture Content.

     Bws  =       Vwc(std)  + Vws9(std)	
      ws     Vwc(std)    wsg(std)     m(std)
     Note:  In saturated or moisture droplet-laden  gas  streams,  two
calculations of the moisture content of the  stack gas  shall  be made,  one
using a value based upon the saturated  conditions (see  Section 1.2),
and another based upon the  results of the impinger  analysis.   The
lower of these two values of B   shall  be considered correct.
                              ws
     2.3.6  Verification of constant sampling rate.  For each  time
increment, determine the AV .   Calculate the average.   If the
                           m
value for any time increment differs from the average  by more than
10 percent, reject the results and repeat the run.
3.  Approximation Method
     The approximation method described below is presented only  as
a suggested method (see Section 1.2).
     3.1  Apparatus.
     3.1.1  Probe.  Stainless steel or glass tubing, sufficiently
heated to prevent water condensation and equipped with  a filter
(either in-stack or heated  out-stack) to remove particulate matter.
A plug of glass wool, inserted into the end of the probe, is a
satisfactory filter.
     3.1.2  Impingers.  Two midget impingers, each with 30 ml
capacity, or equivalent.
     3.1.3  Ice Bath.  Container and ice, to aid in condensing
moisture  in impingers.

-------
                               1-74
     3.1.4  Drying Tube.  Tube packed with new or regenerated 6- to 16-
mesh indicating-type silica gel (or equivalent desiccant), to dry the
sample gas and to protect the meter and pump.
      3.1.5  Valve.  Needle valve, to regulate the sample gas flow rate.
      3.1.6  Pump.  Leak-free, diaphragm type, or equivalent, to pull the
gas  sample through the train.
      3.1.7  Volume meter.  Dry gas meter, sufficiently accurate to
measure the sample volume within 2%, and calibrated over the range of
flow rates and conditions actually encountered during sampling.
      3.1.8  Rate Meter.  Rotameter, to measure the flow range from
0. to 3 1pm (O.to 0.11  cfm).
      3.1.9  Graduated Cylinder.   25 ml.
      3.1.10  Barometer,  Mercury,  aneroid, or other barometer, as
described in Section 2.1.5 above.
      3.1.11  Vacuum Gauge.  At least 760 mm Hg (30 in. Hg) gauge, to be
usad  for the sampling leak check.
      3.2  Procedure.
      3.2.1  Place exactly 5 ml distilled water in each impinger.
Leak check the sampling train as follows:   Temporarily insert a vacuum
gauge at or near the probe inlet;  then, plug the probe inlet and  pull
a vacuum of at least 250 mm Hg (10 in.  Hg).   Note the time rate of
change of the dry gas meter dial;  alternatively,  a  rotameter (0-40
cc/min) may be temporarily attached to the dry gas  meter outlet to
determine the leakage rate.   A leak rate not in excess of 2 percent
of the average sampling rate  is  acceptable.   Note:   Carefully release
the probe inlet plug before turning off the  pump.

-------
                                     1-75
  HEATED PROBE
SILICA GEL TUBE
RATE METER (

    VALVE
FILTER
(GLASS WOOL)
                                                              DRY GAS
                                                             V  METER
    MIDGET IMPINGERS
              PUMP
         Figure 4-4. Moisture-sampling train - approximation method.
    LOCATION.

    TEST
                               COMMENTS
    DATE
    OPERATOR
    BAROMETRIC PRESSURE
CLOCK TIME





GAS VOLUME THROUGH
METER, (Vm),
m3 (ft3)





RATE METER SETTING
m3/min. (ft3/min.)





METER TEMPERATURE.
°C (°F)





      Figure 4-5.  Field moisture determination - approximation method.

-------
                              1-76
    3.2.2  Connect the probe,  insert it into the stack,  and sample
at a constant rate of 2 1pm (0.071  cfm).   Continue sampling until
                                                   o
the dry gas meter registers about 30 liters (1.1 ft )  or until
visible liquid droplets are carried over from the first  impinger
to the second.  Record temperature, pressure, and dry  gas meter
readings as required by Figure 4-5.
     3.2.3  After collecting the sample,  combine the contents of
the two impingers and measure the volume to the nearest  0.5 ml.
     3.3  Calculations.  The calculation method presented is
designed to estimate the moisture in the stack gas; therefore,
other data, which are only necessary for accurate moisture determina-
tions, are not collected.  The following equations adequately
estimate the moisture content, for the purpose of determining
isokinetic sampling rate settings.
     3.3.1  Nomenclature.
     B     = Approximate proportion, by volume, of water vapor in
      wm
             the gas stream leaving the second  impinger, 0.025.
     B     = Water vapor in the gas stream, proportion by volume.
      ws
     M     = Molecular weight of water, 18.0 g/g-mole (18.0 l!b/lb-mole)
     P     = Absolute pressure (for this method, same as barometric
             pressure) at the dry gas meter.
     Pstd  = standard absolute pressure, 760 mm Hg (29.92 in. Hg).
     R     = Ideal gas constant, 0.06236 (m Hq)(m3)/(g-mole)(°K) for
             metric units and 21.85 (in. Hg)(ft3)/lb-mole)(°R) for
             English units.

-------
                                1-77
     T      = Absolute temperature at meter, °K (°R).
      m
     T ..   = Standard absolute temperature, 293°K (528°R).
     Vf     = Final volume of impinger contents, ml.
     V.     = Initial volume of impinger contents, ml.
     V      = Dry gas volume measured by dry gas meter, dcm (dcf).
     V / .,>= Dry gas volume measured by dry gas meter, corrected
              to standard conditions, dscm (dscf).
     V  / .  .\=Volume of water vapor condensed, corrected to
              standard conditions, scm (scf).
     Y      = Dry gas meter calibration factor.
     p ,     = Density of water, 0.9982 g/ml  (0.002201 Ib/ml).
      w
     3.3.2  Volume of water vapor collected.
                  
-------
                                     1-78
     where:
                      o
          K2 = 0.3858  K/mm Hg for metric units.
             = 17.64 °R/in. Hg for English units.
     3.3.4  Approximate moisture content.
                                                 Vwc(std)
                                                          Equation 4-7
4.  Calibration
     4.1  For the reference method,  calibrate equipment as specified in
the following sections of Method 5:   Section 5.3 (metering system);
Section 5.5 (temperature gauges); and Section 5.7 (barometer).   The
recommended leak check of the metering system (Section 5.6 of Method 5)
also applies to the reference method.  For the approximation method, use
the procedures outlined in Section 5.1.1  of Method 6 to calibrate the
metering system, and the procedure of Method 5, Section 5.7 to calibrate
the barometer.
5.  Bibliography
     1.  Air Pollution Engineering Manual (Second Edition). Danielson,
J. A. (ed.).  U. S. Environmental Protection Agency, Office of Air
Quality Planning and Standards.  Research Triangle Park, N. C.  Publication
No. AP-40.  1973.
     2.  Devorkin, Howard, et al.  Air Pollution Source Testing Manual.
Air Pollution Control District, Los Angeles, Calif.  November, 1963.
     3.  Methods for Determination of Velocity, Volume, Dust and Mist
Content of Gases.  Western Precipitation Division of Joy Manufacturing
Co.  Los Angeles, Calif.  Bulletin WP-50.  1968.

-------
DETERMINATION OF PARTICULATE EMISSIONS



       FROM STATIONARY SOURCES





              METHODS

-------
                             1-79
             METHOD 5—DETERMINATION OF PARTICULATE
               EMISSIONS FROM STATIONARY SOURCES
1.  Principle and Applicability
     1.1  Principle.  Particulate matter is withdrawn isokinetically
from the source and collected on a glass fiber filter maintained at
a temperature in the range of 120 +_ 14°C (248 +_25°F) or such other
temperature as specified by an applicable subpart of the standards
or approved by the Administrator, U.  S. Environmental Protection
Agency, for a particular application.  The particulate mass, which
includes any material that condenses  at or above the filtration
temperature, is determined gravimetrically after removal of uncombined
water.
     1.2  Applicability.  This method is applicable for the determina-
tion of particulate emissions from stationary sources.
2.  Apparatus
     2.1  Sampling Train.  A schematic of the sampling train used in
this method is shown in Figure 5-1.  Complete construction details
are given in APTD-0581 (Citation 2 in Section 7); commercial models
of this train are also available.  For changes from APTD-0581 and
for allowable modifications of the train shown in Figure 5-1, see
the following subsections.
     The operating and maintenance procedures for the sampling train
are described in APTD-0576 (Citation  3 in Section 7).  Since correct
usage is important in obtaining valid results, all users should read
APTD-0576 and adopt the operating and maintenance procedures outlined
in it, unless otherwise specified herein.  The sampling train consists
of the following components:

-------
                             1-80
     2.1.1  Probe Nozzle.   Stainless steel  (316)  or glass  with sharp,
tapered leading edge.   The angle of taper shall  be <30°  and the
taper shall be on the  outside to preserve a constant internal  diameter.
The probe nozzle shall be  of the button-hook or  elbow design,  unless
otherwise specified by the Administrator.  If made of stainless steel,
the nozzle shall be constructed from seamless tubing; other materials
of construction may be used, subject to the approval of  the
Administrator.
     A range of nozzle sizes suitable for isokinetic sampling  should
be available, e.g., 0.32 to 1.27 cm (1/8 to 1/2  in.)—or larger if
higher volume sampling trains are used—inside diameter  (ID) nozzles
in increments of 0.16  cm (1/16 in.).  Each nozzle shall  be calibrated
according to the procedures outlined in Section  5.
     2.1.2  Probe Liner.  Borosilicate or quartz glass tubing  with a
heating system capable of maintaining a gas temperature  at the exit
end during sampling of 120 +_ 14°C (248 +_ 25°F),  or such  other  tempera-
ture as specified by an applicable subpart of the standards or
approved by the Administrator for a particular application.  (The
tester may opt to operate the equipment at a temperature lower than
that specified.)  Since the actual temperature at the outlet of the
probe is not usually monitored during sampling,  probes constructed
according to APTD-0581 and utilizing the calibration curves of
APTD-0576 (or calibrated according to the procedure outlined in
APTD-0576) will be considered acceptable.

-------
1-81
                                                o>
                                                c

                                                "o.

                                                E
                                                (D
                                                l/l
                                                O

                                                TO
                                                ro
                                                Q_
                                                in
                                                OJ

-------
                              1-82
     Either borosilicate or quartz glass probe liners  may be used
for stack temperatures up to about 480°C (900°F);  quartz liners
shall be used for temperatures between 480 and 900°C (900 and 1650°F).
Both types of liners may be used at higher temperatures than specified
for short periods of time, subject to the approval  of  the Administrator.
The softening temperature for borosilicate is 820°C (1508°F), and tor
quartz it is 1500°C (2732°F).
     Whenever practical, every effort should be made to use borosilicate
or quartz glass probe liners.  Alternatively, metal liners (e.g., 316
stainless steel, Incoloy 825,  or other corrosion  resistant metals)
made of seamless tubing may be used, subject to the approval of the
Administrator.
     2.1.3  Pi tot Tube.  Type S, as described in Section 2.1 of
Method 2, or other device approved by the Administrator.  The pi tot
tube shall be attached to the probe (as shown in Figure 5-1) to allow
constant monitoring of the stack gas velocity.  The impact (high
pressure) opening plane of the pitot tube shall be even with or
above the nozzle entry plane (see Method 2, Figure 2-6b) during
sampling.  The Type S pitot tube assembly shall have a known coefficient,
determined as outlined in Section 4 of Method 2.
     2.1.4  Differential Pressure Gauge.  Inclined manometer or
equivalent device (two), as described in Section 2.2 of Method 2.
One manometer shall be used for velocity head (AP) readings, and
the other, for orifice differential pressure readings.
 Mention of trade names or specific products does not constitute
 endorsement by the Environmental Protection Agency.

-------
                              1-83
     2.1.5  Filter Holder.   Borosllicate glass,  with a glass  frit
filter support and a silicone rubber gasket.   Other materials of
construction (e.g., stainless steel, Teflon,  Viton) may be used,
subject to the approval  of the Administrator.   The holder design shall
provide a positive seal  against leakage from  the outside or around the
filter.  The holder shall be attached immediately at the outlet of
the probe (or cyclone, if used).
     2.1.6  Filter Heating System.   Any heating  system capable of
maintaining a temperature around the filter holder during sampling
of 120 +_ 14°C (248 + 25°F), or such other temperature as specified by
an applicable subpart of the standards or approved by the
Administrator for a particular application.  Alternatively, the tester
                                             /
may opt to operate the equipment at a temperature lower than  that
specified.  A temperature gauge capable of measuring temperature to
within 3°C (5.4°F) shall be installed so that the temperature around
the filter holder can be regulated and monitored during sampling.
Heating systems other than the one shown in APTD-0581 may be  used.
     2.1.7  Condenser.  The following system shall be used to determine
the stack gas moisture content:  Four impingers  connected in  series
with leak-free ground glass fittings or any similar leak-free non-
contaminating fittings.   The first, third, and fourth impingers shall
be of the Greenburg-Smith design, modified by replacing the tip with
a 1.3 cm (1/2 in.) ID glass tube extending to about 1.3 cm (1/2 in.)
from the bottom of the flask.  The second impinger shall be of the
Greenburg-Smith design with the standard tip.   Modifications  (e.g.,

-------
                             1-84
using flexible connections between  the impingers, using materials
other than glass, or using flexible vacuum lines to connect the
filter holder to the condenser) may be used, subject to the
approval of the Administrator.   The first and second impingers shall
contain known quantities of water (Section 4.1.3), the third shall
be empty, and the fourth shall  contain a known weight of silica gel,
or equivalent desiccant.  A thermometer, capable of measuring tempera-
ture to within 1°C (2°F) shall  be placed at the outlet of the fourth
impinger for monitoring purposes.
     Alternatively, any system that cools the sample gas stream and
allows measurement of the water condensed and moisture 'leaving the
condenser, each to within 1 ml  or 1 g may be used, subject to the
approval of the Administrator.   Acceptable means are to measure the
condensed water either gravimetrically or volumetrically and to measure
the moisture leaving the condenser by:  (1) monitoring the temperature
and pressure at the exit of the condenser and using Dal ton's law of
partial pressures; or (2) passing the sample gas stream through a
tared silica gel (or equivalent desiccant) trap with exit gases kept
below 20°C (68°F) and determining the weight gain.
     If means other than silica gel are used to determine the amount  of
moisture leaving the condenser, it is recommended that silica gel (or
equivalent) still be used between the condenser system and pump to
prevent moisture condensation in the pump and metering devices and
to avoid the need to make corrections for moisture in the metered
volume.

-------
                              1-85
     Note:  If a determination of the participate matter collected
in the impingers is desired in addition to moisture content,  the
impinger system described above shall be used, without.modification.
Individual States or control agencies requiring this information
shall be contacted as to the sample recovery and analysis of  the
impinger contents.
     2.1.8  Metering System.  Vacuum gauge, leak-free pump, thermometers
capable of measuring temperature to within 3°C (5.4°F), dry gas meter
capable of measuring volume to within 2 percent, and related  equipment,
as shown in Figure 5-1.  Other metering systems capable of maintaining
sampling rates within 10 percent of isokinetic and of determining
sample volumes to within 2 percent may be used, subject to the approval
of the Administrator.  When the metering system is used in conjunction
with a pitot tube, the system shall enable checks of isokinetic rates.
     Sampling trains utilizing metering systems designed for  higher
flow rates than that described in APTD-0581 or APTD-0576 may  be used
provided that the specifications of this method are met.
     2.1.9  Barometer.  Mercury, aneroid, or other barometer  capable
of measuring atmospheric pressure to within 2.5 mm Hg (0.1 in. Hg).
In many cases, the barometric reading may be obtained from a  nearby
national weather service station, in which case the station value
(which is the absolute barometric pressure) shall be requested and
an adjustment for elevation differences between the weather station
and sampling point shall be applied at a rate of minus 2.5 mm Hg
(0.1 in. Hg) per 30 m  (100 ft) elevation increase or vice versa
for elevation decrease.

-------
                             1-86
     2.1.10  Gas Density Determination Equipment.   Temperature
sensor and pressure gauge, as described in Sections 2.3 and  2.4 of
Method 2, and gas analyzer, if necessary,  as  described  in  Method 3.
The temperature sensor shall, preferably,  be  permanently attached
to the pi tot tube or sampling probe in a fixed configuration,  such
that the tip of the sensor extends beyond  the leading edge of  the
probe sheath and does not touch any metal. Alternatively, the sensor
may be attched just prior to use in the field.  Note, however, that
if the temperature sensor is attached in the  field, the sensor must
be placed in an interference-free arrangement with respect to  the
Type S pi tot tube openings (see Method 2,  Figure 2-7).   As a second
alternative, if a difference of not more than 1  percent in the average
velocity measurement is to be introduced,  the temperature  gauge need
not be attached to the probe or pitot tube.  (This alternative is
subject to the approval of the Administrator.)
     2.2  Sample Recovery.  The following items are needed:
     2.2.1  Probe-Liner and Probe-Nozzle Brushes.   Nylon bristle
brushes with stainless steel wire handles. The probe brush  shall
have extensions (at least as long as the probe) of stainless steel,
Nylon, Teflon, or similarly inert material.  The brushes shall be
properly sized and shaped to brush out the probe liner  and nozzle.
     2.2.2  Wash Bottles—Two.  Glass wash bottles are  recommended;
polyethylene wash bottles may be used at the  option of  the tester.
It is recommended that acetone not be stored  in polyethylene bottles
for longer than a month.

-------
                               1-87
     2.2.3  Glass Sample Storage Containers.   Chemically resistant,
borosilicate glass bottles, for acetone washes, 500 ml  or 1000 ml.
Screw cap liners shall  either be rubber-backed Teflon or shall be
constructed so as to be leak-free and resistant to chemical  attack
by acetone.  (Narrow mouth glass bottles have been found to  be less
prone to leakage.)  Alternatively, polyethylene bottles may  be used.
     2.2.4  Petri Dishes.  For filter samples, glass or polyethylene,
unless otherwise specified by the Administrator.
     2.2.5  Graduated Cylinder and/or Balance.  To measure condensed
water to within 1 ml or 1 g.  Graduated cylinders shall have sub-
divisions no greater than 2 ml.  Most laboratory balances are capable
of weighing to the nearest 0.5 g or less.  Any of these balances is
suitable for use here and in Section 2.3.4.
     2.2.6  Plastic Storage Containers.  Air-tight containers to
store silica gel.
     2.2.7  Funnel and Rubber Policeman.  To aid in transfer of silica
gel to container; not necessary if silica gel 1s weighed 1n  the field.
     2.2.8  Funnel.  Glass or polyethylene, to aid in sample recovery.
     2.3  Analysis.  For analysis, the following equipment is needed:
     2.3.1  Glass Weighing Dishes.
     2.3.2  Desiccator.
     2.3.3  Analytical Balance.  To measure to within 0.1 mg.
     2.3.4  Balance.  To measure to within 0.5 g.
     2.3.5  Beakers.  250 ml.
     2.3.6  Hygrometer.  To measure the relative humidity of the
laboratory environment.

-------
                              1-88
     2.3.7  Temperature Gauge.   To measure the temperature  of the
laboratory environment.
3.  Reagents
     3.1  Sampling.   The reagents used in sampling are as follows:
     3.1.1  Filters.   Glass fiber filters, without organic  binder,
exhibiting at least 99.95 percent efficiency (<0.05 percent penetration)
on 0.3-micron dioctyl phthalate  smoke particles.   The filter efficiency
test shall be conducted in accordance with ASTM standard method
D 2986-71.  Test data from the  supplier's quality control program are
sufficient for this purpose. .
     3.1.2  Silica Gel.  Indicating.type, 6 to 16 mesh.   If previously
used, dry at 175°C (350°F) for  2 hours.   New silica gel  may be used
as received.  Alternatively, other types of desiccants (equivalent or
better) may be used, subject to  the approval of the Administrator.
     3.1.3  Water.  When analysis of the material caught in the
impingers is required, distilled water shall be used.  Run  blanks
prior to field use to eliminate  a high blank on test samples.
     3.1.4  Crushed Ice.
     3.1.5  Stopcock Grease. Acetone-insoluble,  heat-stable silicone
grease.  This is not necessary  if screw-on connectors with  Teflon
sleeves, or similar, are used.   Alternatively, other types  of stopcock
grease may be used, subject to  the approval of the Administrator.
     3.2  Sample Recovery.  Acetone—reagent grade, ^0.001  percent
residue, in glass bottles—is required.   Acetone  from metal containers
generally has a high residue blank and should not be used.   Sometimes,

-------
                              1-89
suppliers transfer acetone to glass bottles from metal  containers;
this, acetone blanks shall be run prior to field use and only
acetone with low blank values (<0.001 percent) shall be used.  In
no case shall a blank value of greater than 0.001 percent of the
weight of acetone used be subtracted from the sample weight.
     3.3  Analysis.  Two reagents are required for the analysis:
     3.3.1  Acetone.  Same as 3.2.
     3.3.2  Desiccant.  Anhydrous calcium suTfate, indicating type.
Alternatively, other types of desiccants may be used, subject to the
approval of the Administrator.
4.  Procedure
     4.1  Sampling.  The complexity of this method is such that, in
order to obtain reliable results, testers should be trained and
experienced with the test procedures.
     4.1.1  Pretest Preparation.  All the components shall be maintained
and calibrated according to the procedure described in APTD-0576, unless
otherwise specified herein.
     Weigh several 200 to 300 g portions of silica gel in air-tight
containers to the nearest 0.5 g.  Record the total weight of the
silica gel plus container, on each container.  As an alternative, the
silica gel need not be preweighed, but may be weighed directly in its
impinger or sampling holder just prior to train assembly.
     Check filters visually against  light for irregularities and
flaws or pinhole leaks.   Label filters of the proper diameter on the
back side near the edge using numbering machine ink.  As an alternative,

-------
                                1-90
label the shipping containers (glass or plastic petri dishes) and
keep the filters in these containers at all  times except during
sampling and weighing.
     Desiccate the filters at 20 +_ 5.6°C (68 +_ 10°F) and ambient
pressure for at least 24 hours and weigh at intervals of at least
6 hours to a constant weight, i.e., <0.5 mg change from previous
weighing; record results to the nearest 0.1  mg.  During each
weighing the filter must not be exposed to the laboratory atmosphere
for a period greater than 2 minutes and a relative humidity above
50 percent.  Alternatively (unless otherwise specified by the
Administrator), the filters may be oven dried at 105°C (220°F) for
2 to 3 hours, desiccated for 2 hours, and weighed.  Procedures other
than those described, which account for relative humidity effects,
may be used, subject to the approval of the Administrator.
     4.1.2  Preliminary Determinations.  Select the sampling site and
the minimum number of sampling points according to Method 1 or as
specified by the Administrator.  Determine the stack pressure,
temperature, and the range of velocity heads using Method 2; it is
recommended that a leak-check of the pi tot lines (see Method 2,
Section 3.1) be performed.  Determine the moisture content using
Approximation Method 4 or its alternatives for the purpose of making
isokinetic sampling rate settings.  Determine the stack gas dry
molecular weight, as described in Method 2, Section 3.6; if integrated
Method 3 sampling is used for molecular weight determination, the
Integrated bag sample shall be taken simultaneously with, and for
the same total length of time as, the particulate sample run.

-------
                                 1-91
     Select a nozzle size based on the range of velocity heads, such
that it is not necessary to change the nozzle size in order to maintain
isokinetic sampling rates.  During the run, do not change the nozzle
size.  Ensure that the proper differential pressure gauge is chosen for
the range of velocity heads encountered (see Section 2.2 of Method 2).
     Select a suitable probe liner and probe length such that all
traverse points can be sampled.  For large stacks, consider sampling
from opposite sides of the stack to reduce the length of probes.
     Select a total sampling time greater than or equal  to the minimum
total sampling time specified in the test procedures for the specific
industry such that (1) the sampling time per point is not less than 2
min.(or some greater time interval as specified by the Administrator),
and (2) the sample volume taken (corrected to standard conditions) will
exceed the required minimum total gas sample volume.  The latter is
based on an approximate average sampling rate.
     The sampling time at each point shall be the same.   It is recom-
mended that the number of minutes sampled at each point  be an integer or
an integer plus onehalf minute, in order to avoid timekeeping errors.
     In some circumstances, e.g., batch cycles, it may be necessary to
sample for shorter times at the traverse points and to obtain smaller
gas sample volumes.  In these cases, the Administrator's approval  must
first be obtained.
     4.1.3  Preparation of Collection Train.   During preparation and
assembly of the sampling train, keep all openings where  contamination

-------
                               1-92
can occur covered until  just prior to assembly or until  sampling
is about to begin.
     Place 100 ml of water in each of the first two irnpingers, leave
the third impinger empty, and transfer approximately 200 to 300 g of
preweighed silica gel from its container to the fourth impinger.
More silica gel may be used, but care should be taken to ensure that
it is not entrained and carried out from the impinger during sampling.
Place the container in a clean place for later use in the sample
recovery.  Alternatively, the weight of the silica gel plus impinger
may be determined to the nearest 0.5 g and recorded.
     Using a tweezer or clean disposable surgical gloves, place a
labeled  (identified) and weighed filter in the filter holder.  Be sure
that the filter is properly centered and the gasket properly placed
so as to prevent the sample gas stream from circumventing the filter.
Check the filter for tears after assembly is completed.
     When glass liners are used, install the selected nozzle using
a Viton  A 0-ring when stack temperatures are less than 260°C (500°F)
and an asbestos string gasket when temperatures are higher.  See
APTD-0576 for details.  Other connecting systems using either 316
stainless steel or Teflon ferrules may be used.  When metal liners
are used, install the nozzle as above or by a leak-free direct
mechanical connection.  Mark the probe with heat resistant tape or
by some  other method to denote the proper distance Into the stack or
duct for each sampling point.

-------
                              1-93
     Set up the train as in Figure 5-1, using (if necessary) a very
light coat of silicone grease on all  ground glass joints, greasing
only the outer portion (see APTD-0576) to avoid possibility of
contamination by the silicone grease.   Subject to the approval of
the Administrator, a glass cyclone may be used between the probe and
filter holder when the total particulate catch is expected to exceed
100 mg or when water droplets are present in the stack gas.
     Place crushed ice around the impingers.
     4.1.4  Leak-Check Procedures.
     4.1.4.1  Pretest Leak-Check.  A pretest leak-check is recommended,
but not required.  If the tester opts  to conduct the pretest leak-check,
the following procedure shall be used.
     After the sampling train has been assembled, turn on and set the
filter and probe heating systems at the desired operating temperatures.
Allow time for the temperatures to stabilize.  If a Viton A 0-ring or
other leak-free connection is used in  assembling the probe nozzle to
the probe liner, leak-check the train  at the sampling site by plugging
the nozzle and pulling a 380 mm Hg (15 in.  Hg) vacuum.
     Note:  A lower vacuum may be used, provided that it is not exceeded
during the test.
     If an asbestos string is used, do not connect the probe to the
train during the leak-check.  Instead, leak-check the train by first
plugging the inlet to the filter holder (cyclone, if applicable) and
pulling a 380 mm Hg (15 in. Hg) vacuum (see Note Immediately above).
Then connect the probe to the train and leak-check at about 25 mm Hg
(1 in. Hg) vacuum; alternatively, the  probe may be leak-checked with

-------
                               1-94
the rest of the sampling train, in one step, at 380 mm Hg (15 in.  Hg)
vacuum.  Leakage rates in excess of 4 percent of the average sampling
                 ^
rate or 0.00057 m /min (0.02 cfm), whichever is less, are unacceptable.
     The following leak-check instructions for the sampling train
described in APTD-0576 and APTD-0581 may be helpful.  Start the pump
with bypass valve fully open and coarse adjust valve completely closed.
Partially open the coarse adjust valve and slowly close  the bypass
valve until the desired vacuum is reached.  Do not reverse direction
of bypass valve; this will cause water to back up into the filter
holder.  If the desired vacuum is exceeded, either leak-check at
this higher vacuum or end the leak check as shown below and start  over.
     When the leak-check is completed, first slowly remove the plug
from the inlet to the probe, filter holder, or cyclone (if applicable)
and immediately turn off the vacuum pump.  This prevents the water in
the impingers from being forced backward into the filter holder and
silica gel from being entrained backward into the third impinger.
     4.1.4.2  Leak-Checks During Sample Run.  If, during the sampling
run, a component  (e.g., filter assembly or impinger) change becomes
necessary, a leak-check shall be conducted immediately before the
change is made.  The leak-check shall be done according to the procedure
outlined in Section 4.1.4.1 above, except that it shall be done at a
vacuum equal to or greater than the maximum value recorded up to that
point in the test.  If the leakage rate is found to be no greater  than
         o
0.00057 m /min  (0.02 cfm) or 4 percent of the average sampling rate

-------
                             1-95
(whichever is less), the results are acceptable, and no correction
will need to be applied to the total volume of dry gas metered;
if, however, a higher leakage rate is obtained, the tester shall
either record the leakage rate and plan to correct the sample volume
as shown in Section 6.3 of this method, or shall void the sampling run.
     Immediately after component changes, leak-checks are optional;
if such leak-checks are done, the procedure outlined in Section  4.1.4.1
above shall be used.
     4.1.4.3  Post-test Leak-Check.  A leak-check is mandatory at the
conclusion of each sampling run.  The leak-check shall be done in
accordance with the procedures outlined in Section 4.1.4.1, except
that it shall be conducted at a vacuum equal to or greater than  the
maximum value reached during the sampling run.  If the leakage rate
is found to be no greater than 0.00057 m /min (0.02 cfm) or 4 percent
of the average sampling rate (whichever is less), the results are
acceptable, and no correction need be applied to the total volume of
dry gas metered.  If, however, a higher leakage rate is obtained, the
tester shall either record the leakage rate and correct the sample
volume as shown in Section 6.3 of this method, or shall void the
sampling run.
     4.1.5  Particulate Train Operation.  During the sampling run,
maintain an isokinetic sampling rate (within 10 percent of true
isokinetic unless otherwise specified by the Administrator) and a
temperature around the filter of 120 +_ 14°C (248 + 25°F), or such other

-------
                              1-96
temperature as specified by an applicable subpart of the standards
or approved by the Administrator.
     For each run, record the data required on a data sheet such as
the one shown in Figure 5-2.  Be sure to record the initial dry gas
meter reading.  Record the dry gas meter readings at the beginning
and end of each sampling time increment, when changes in flow rates
are made, before and after each leak check, and when sampling is halted.
Take other readings required by Figure 5-2 at least once at each sample
point during each time increment and additional readings when significant
changes (20 percent variation in velocity head readings) necessitate
additional adjustments in flow rate.  Level and zero the manometer.
Because the manometer level and zero may drift due to vibrations and
temperature changes, make periodic checks during the traverse.
     Clean the portholes prior to the test run to minimize the chance
of sampling deposited material.  To begin sampling, remove the nozzle
cap, verify that the filter and probe heating systems are up to
temperature, and that the pitot tube and probe are properly positioned.
Position the nozzle at the first traverse point with the tip pointing
directly into the gas stream.  Immediately start the pump and adjust
the flow to isokinetic conditions.  Nomographs are available, which
aid in the rapid adjustment of the isokinetic sampling rate without
excessive computations.  These nomographs are designed for use when the
Type S pitot tube coefficient is 0.85 +^0.02, and the stack gas
equivalent density (dry molecular weight) is equal to 29 +_4.  APTD-0576
details the procedure for using the nomographs.  If C  and M. are
outside the above stated ranges, do not use the nomographs unless
appropriate steps (see Citation 7 in Section 7) are taken to compensate
for the deviations.

-------
                          1-97






























i^ "j'
a OC ^



















c
"i
u
CC
uu
LU
-  "J £ «s t
< CO CC 0 2
£ S ? E- = £
Q. Q. CO X (_ ~
^ if o J^ z *^
t- DC 5 z Q °
. •- Q LU — UJ
^ "J UJ — ' UJ «
uj 2 5 uj — ' 2
— o — . CD rsi cc
00 cc ^ a M "J
S < CO CC C3 >
 "•
(J
O
cc
U
a£
O
I
O





Z
0





cc
0






— t—
*—  **
e>
!i o
o u
0

*— o
UJ 	
Z (J
O

UJ
Q. UJ —
5 5
O
_l
K ^ "> UJ
= u? < O 5
a. u.
D
^
O O
E c

O
5 9 i;>~
2*5; E
UJ ^
U.
cc
<2
£;|
UJ
- i

'•^XL
— o
o

i*f
c_
>
O
Z
a.





i I
c
h- —
Z
O a:
a. uj
UJ °°
lo 5
K

























































































































































































































































































































































































J


C>
<



















<
0







Ol
*


















UJ
O

-------
                              1-98
     When the stack is under significant negative pressure (height
of impinger stem), take care to close the coarse adjust valve before
inserting the probe into the stack to prevent water from backing into
the filter holder.  If necessary, the pump may be turned on with the
coarse adjust valve closed.
     When the probe is in position, block off the openings around
the probe and porthole to prevent unrepresentative dilution of the
gas stream.
     Traverse the stack cross-section, as required by Method 1 or as
specified by the Administrator, being careful not to bump the probe
nozzle into the stack walls  when sampling near the walls or when
removing or inserting the probe through the portholes; this minimizes
the chance of extracting deposited material.
     During the test run, make periodic adjustments to keep the
temperature around the filter holder at the proper level; add more
ice and, if necessary, salt  to maintain a temperature of less than
20°C (68°F) at the condenser/silica gel outlet.  Also, periodically
check the level and zero of the manometer.
     If the pressure drop across the filter becomes too high, making
isokinetic sampling difficult to maintain, the filter may be replaced
in the midst of a sample run.  It is recommended that another complete
filter assembly be used rather than attempting to change the filter
itself.  Before a new filter assembly is installed, conduct a leak-check
(see Section 4.1.4.2).  The total particulate weight shall include the
summation of all filter assembly catches.

-------
                             1-99
     A single train shall be used for the entire sample run,  except
in cases where simultaneous sampling is required in two or more
separate ducts or at two or more different locations within the same
duct, or, in cases where equipment failure necessitates a change of
trains.   In all other situations, the use of two or more trains will
be subject to the approval of the Administrator.
     Note that when two or more trains are used, separate analyses of
the front-half and (if applicable) impinger catches from each train
shall be performed, unless identical nozzle sizes were used on all
trains, in which case, the front-half catches from the individual trains
may be combined (as may the impinger catches) and one analysis of front-
half catch and one analysis of impinger catch may be performed.  Consult
with the Administrator for details concerning the calculation of
results when two or more trains are used.
     At the end of the sample run, turn off the coarse adjust valve,
remove the probe and nozzle from the stack, turn off the pump, record
the final dry gas meter reading, and conduct a post-test leak-check, as
outlined in Section 4.1.4.3.  Also, leak-check the pitot lines as
described in Method 2, Section 3.1; the lines must pass this leak-check,
in order to validate the velocity head data.
     4.1.6  Calculation of Percent Isokinetic.  Calculate percent
isokinetic  (see Calculations, Section 6) to determine whether the run
was valid or another test run should be made.  If there was difficulty
in maintaining isokinetic rates due to source conditions, consult with
the Administrator for possible variance on the isokinetic rates.

-------
                              I-100
     4.2  Sample Recovery.   Proper cleanup procedure begins as soon
as the probe is removed from the stack at the end of the sampling
period.  Allow the probe to cool.
     When the probe can be  safely handled, wipe off all  external
particulate matter near the tip  of the probe nozzle and  place a cap
over it to prevent losing or gaining particulate matter.  Do not cap
off the probe tip tightly while  the sampling train is cooling down
as this would create a vacuum in the filter holder, thus drawing water
from the impingers into the filter holder.
     Before moving the sample train to the cleanup site, remove the
probe from the sample train, wipe off the silicone grease, and cap
the open outlet of the probe.  Be careful not to lose any condensate
that might be present.  Wipe off the silicone grease from the filter
inlet where the probe was fastened and cap it.  Remove the umbilical
cord from the last impinger and  cap the impinger.  If a  flexible line
is used between the first impinger or condenser and the  filter holder,
disconnect the line at the filter holder and let any condensed water
or liquid drain into the impingers or condenser.  After  wiping off the
silicone grease, cap off the filter holder outlet and impinger inlet.
Either ground-glass stoppers, plastic caps, or serum caps may be used
to close these openings.
     Transfer the probe and filter-impinger assembly to  the cleanup
area.  This area should be clean and protected from the  wind so that
the chances of contaminating or  losing the sample will be minimized.

-------
                                    1-101
     Save a portion of the acetone used for cleanup as a blank.   Take
200 ml of this acetone directly from the wash bottle being used  and
place it in a glass sample container labeled "acetone blank."
     Inspect the train prior to and during disassembly and note  any
abnormal conditions.  Treat the samples as follows:
     Container No. 1.  Carefully remove the filter from the filter holder
and place it in its identified petri dish container.  Use a pair of
tweezers and/or clean disposable surgical gloves to handle the filter.
If it is necessary to fold the filter, do so such that the particulate
cake is inside the fold.  Carefully transfer to the petri dish any
particulate matter and/or filter fibers which adhere to the filter
holder gasket, by using a dry Nylon bristle brush and/or a sharp-edged
blade.  Seal the container.
     Container No. 2.  Taking care to see that dust on the outside
of the probe or other exterior surfaces does not get into the sample,
quantitatively recover particulate matter or any condensate from the
probe nozzle, probe fitting, probe liner, and front half of the
filter holder by washing these components with acetone and placing
the wash in a glass container.  Distilled water may be used instead
of acetone when approved by the Administrator and shall be used when
specified by the Administrator; in these cases, save a water blank
and follow the Administrator's directions on analysis.  Perform the
acetone rinses as follows:

-------
                                  1-102
     Carefully remove the probe nozzle and clean the inside surface
by rinsing with acetone from a wash bottle and brushing with a Nylon
bristle brush.  Brush until the acetone rinse shows no visible particles,
after which make a final rinse of the inside surface with acetone.  ,
     Brush and rinse the inside parts of the Swagelok fitting with
acetone in a similar way until no visible particles remain.
     Rinse the probe liner with acetone by tilting and rotating the
probe while squirting acetone into its upper end so that all inside
surfaces will be wetted with acetone.  Let the acetone drain from the
lower end into the sample container.  A funnel (glass or polyethylene)
may be used to aid in transferring liquid washes to the container.   Follow
the acetone rinse with a probe brush.  Hold the probe in an inclined
position, squirt acetone into the upper end as the probe brush is being
pushed with a twisting action through the probe; hold a sample container
underneath the lower end of the probe, and catch any acetone and particu-
late matter which is brushed from the probe.  Run the brush through the
probe three times or more until no visible particulate matter is carried
out with the acetone or until none remains in the probe liner on visual
inspection.  With stainless steel or other metal probes, run the brush
through in the above prescribed manner at least six times since metal
probes have small crevices in which particulate matter can be entrapped.
Rinse the brush with acetone, and quantitatively collect these washings
in the sample container.  After the brushing, make a final acetone rinse
of the probe  as described above.

-------
                                 1-103
     It is recommended that two people be used to clean the probe
to minimize sample losses.   Between sampling runs, keep brushes clean
and protected from contamination.
     After ensuring that all joints have been wiped clean of silicone
grease, clean the inside of the front half of the filter holder by
rubbing the surfaces with a Nylon bristle brush and rinsing with
acetone.  Rinse each surface three times or more if needed to remove
visible particulate.  Make a final rinse of the brush and filter
holder.  Carefully rinse out the glass cyclone, also (if applicable).
After all acetone washings and particulate matter have been collected
in the sample container, tighten the lid on the sample container so
that acetone will not leak out when it is shipped to the laboratory.
Mark the height of the fluid level to determine whether or not
leakage occurred during transport.  Label the container to clearly
identify its contents.
     Container No. 3.  Note the color of the indicating silica gel
to determine if it has been completely spent and make a notation of
its condition.  Transfer the silica gel from the fourth impinger to
its original container and  seal.  A funnel may make it easier to pour
the silica gel without spilling.  A rubber policeman may be used as
an aid  in removing the silica gel from the impinger.  It is not
necessary to remove the small amount of dust particles that may adhere
to the  impinger wall and are difficult to remove.  Since the gain in
weight  is to be used for moisture calculations, do not use any water

-------
                            1-104
or other liquids to transfer the silica gel.   If a balance is
available in the field, follow the procedure  for container No.  3
in Section 4.3.
     Impinger Water.  Treat the impingers as  follows:   Make a
notation of any color or film in the liquid catch.  Measure the
liquid which is in the first three impingers  to within +1  ml  by
using a graduated cylinder or by weighing it  to within +0.5 g by
using a balance (if one is available).   Record the volume  or weight
of liquid present.  This information is required to calculate the
moisture content of the effluent gas.
     Discard the liquid after measuring and recording  the  volume or
weight, unless analysis of the impinger catch is required  (see Note,
Section 2.1.7).
     If a different type of condenser is used, measure the amount of
moisture condensed either volumetrically or gravimetrically.
     Whenever possible, containers should be  shipped in such a way that
they remain upright at all times.
     4.3  Analysis.  Record the data required on a sheet such as the
one shown in Figure 5-3.  Handle each sample  container as  follows:
     Container No. 1.  Leave the contents in  the shipping  container
or transfer the filter and any loose particulate from the  sample
container to a tared glass weighing dish.  Desiccate for 24 hours
in a desiccator containing anhydrous calcium  sulfate.   Weigh to a
constant weight and report the results to the nearest 0.1  mg.  For
purposes of this Section, 4.3, the term "constant weight"  means

-------
                               1-105
Plant	
Date	
Run No	
Filter No	
Amount liquid lost during transport
Acetone blank volume, ml	
Acetone wash volume, ml	
Acetone blank concentration, mg/mg (equation 5-4).
Acetone wash blank, mg (equation 5-5)	
CONTAINER
NUMBER
1
2
TOTAL
WEIGHT OF PARTICULATE COLLECTED,
mg
FINAL WEIGHT


^X^
TARE WEIGHT


^XI^
Less acetone blank
Weight of paniculate matter
WEIGHT GAIN






FINAL
INITIAL
LIQUID COLLECTED
TOTAL VOLUME COLLECTED
VOLUME OF LIQUID
WATER COLLECTED
IMPINGER
VOLUME.
ml




SILICA GEL
WEIGHT.
9



9* ml
     'CONVERT WEIGHT OF WATER TO VOLUME BY DIVIDING TOTAL WEIGHT
      INCREASE BY DENSITY OF WATER (1g/ml).
                                  INCREASE, g
                                     1 g/ml
                                              = VOLUME WATER, ml
                         Figure 5-3. Analytical data.

-------
                               1-106
a difference of no more than 0.5 mg or 1  percent of total  weight
less tare weight, whichever is greater, between two consecutive
weighings, with no less than 6 hours of desiccation time between
weighings.
     Alternatively, the sample may be oven dried at 105°C (220°F) for
2 to 3 hours, cooled in the desiccator, and weighed to a constant
weight, unless otherwise specified by the Administrator.  The tester
may also opt to oven dry the sample at 105°C (220°F) for 2 to 3 hours,
weigh the sample, and use this weight as  a final weight.
     Container No. 2.  Note the level of liquid in the container and
confirm on the analysis sheet whether or not leakage occurred during
transport.  If a noticeable amount of leakage has occurred, either void
the sample or use methods, subject to the approval of the Administrator,
to correct the final results.  Measure the liquid in this container
either volumetrically to +1 ml or gravimetrically to +0.5 g.  Transfer
the contents to a tared 250-ml beaker and evaporate to dryness at ambient
temperature and pressure.  Desiccate for 24 hours and weigh to a
constant weight.  Report the results to the nearest 0.1 mg.
     Container No. 3.  Weigh the spent silica gel (or silica gel plus
impinger) to the nearest 0.5 g using a balance.  This step may be con-
ducted in the field.
     "Acetone Blank" Container.  Measure acetone in this container
either volumetrically or gravimetrically.  Transfer the acetone to
a tared 250-ml beaker and evaporate to dryness at ambient temperature
and pressure.  Desiccate for 24 hours and weigh to a constant weight.
Report the results to the nearest 0.1 mg.

-------
                               1-107
     Note:  At the option of the tester, the contents of Container
No. 2 as well as the acetone blank container may be evaporated at
temperatures higher than ambient.  If evaporation is done at an
elevated temperature, the temperature must be below the boiling point
of the solvent; also, to prevent "bumping," the evaporation process
must be closely supervised, and the contents of the beaker must be
swirled occasionally to maintain an even temperature.  Use extreme
care, as acetone is highly flammable and has a low flash point.
5.  Calibration
     Maintain a laboratory log of all calibrations.
     5.1  Probe Nozzle.  Probe nozzles shall be calibrated before
their initial use in the field.  Using a micrometer, measure the
inside diameter of the nozzle to the nearest 0.025 mm (0.001 in.).
Make three separate measurements using different diameters each time,
and obtain the average of the measurements.  The difference between
the high and low numbers shall not exceed 0.1 mm (0.004 in.).  When
nozzles become nicked, dented, or corroded, they shall  be reshaped,
sharpened, and recalibrated before use.  Each nozzle shall be per-
manently and uniquely identified.
     5.2  Pitot Tube.  The Type S pitot tube assembly shall be calibrated
according to the procedure outlined in Section 4 of Method 2.
     5.3  Metering System.  Before its initial use in the field, the
metering system shall be calibrated according to the procedure outlined
in APTD-0576.  Instead of physically adjusting the dry  gas meter dial
readings to correspond to the v/et test meter readings,  calibration
factors may be used to mathematically correct the gas meter dial readings
to the proper values. Before  calibrating  the-metering system,  it  is  sug-
gested that a leak-check HP conducted.  For metering systems having diaphragm

-------
                              1-108
pumps, the normal leak-check procedure will not detect leakages within
the pump.  For these cases the following leak-check procedure is
suggested: make a 10-minute calibration run at 0.00057 m3/min (0.02 cfm);
at the end of the run, take the difference of the measured wet test meter
and dry gas meter volumes; divide the difference by 10, to get the leak
rate.  The leak rate should not exceed 0.00057 m3/min (0.02 cfm).
     After each field use, the calibration of the metering system
shall be checked by performing three calibration runs at a single,
intermediate orifice setting (based on the previous field test), with
the vacuum set at the maximum value reached during the test series.
To adjust the vacuum, insert a valve between the wet test meter and
the inlet of the metering system.  Calculate the average value of the
calibration factor.  If the calibration has changed by more than 5 per-
cent, recalibrate the meter over the full  range of orifice settings, as
outlined in APTD-0576.
     Alternative procedures, e.g., using the orifice meter coeffi-
cients, may be used, subject to the approval of the Administrator.
     Note:  If the dry gas meter coefficient values obtained before
and after a test series differ by more than 5 percent, the test
series shall either be voided, or calculations for the test series
shall be performed using whichever meter coefficient value (i.e.,
before or after) gives the lower value of total  sample volume.
     5.4  Probe Heater Calibration.   The probe heating system shall be
calibrated before its initial  use in the field according to the pro-
cedure outlined in APTD-0576.   Probes constructed according to APTD-0581
need not be calibrated if the  calibration  curves in APTD-0576 are  used.
     5.5  Temperature Gauges.   Use the procedure in Section 4.3 of
Method 2 to calibrate in-stack temperature gauges.   Dial thermometers,

-------
                             1-109



such as are used for the dry gas meter and condenser outlet, shall  be



calibrated against mercury-in-glass  thermometers.



     5.6  Leak Check of Metering System Shown in Figure 5-1.  That



portion of the sampling train from the pump to the orifice meter



should be leak checked prior to initial use and after each shipment.



Leakage after the pump will result in less volume  being recorded than



is actually sampled.  The following  procedure is suggested (see



Figure 5-4):  Close the main valve on the meter box.  Insert a one-



hole rubber stopper with rubber tubing attached into the orifice



exhaust pipe.  Disconnect and vent the low side of the orifice manometer.



Close off the low side orifice tap.   Pressurize the system to 13 to



18 cm (5 to 7 in.) water column by blowing into the rubber tubing.



Pinch off the tubing and observe the manometer for one minute.  A loss



of pressure on the manometer indicates a leak in the meter box; leaks,



if present, must be corrected.



     5.7  Barometer.  Calibrate against a mercury  barometer.



6.  Calculations



     Carry out calculations, retaining at least one extra decimal



figure beyond that of the acquired data.  Round off figures after



the final calculation.  Other forms  of the equations may be used as



long as they give equivalent results.



     6.1  Nomenclature.


                                               2    2
     A      = Cross-sectional area of nozzle, m  (ft ).



     B      = Water vapor in the gas stream, proportion by volume.
      W o


     C      = Acetone blank residue concentration, mg/g.
      G


     c      = Concentration of particulate matter  in stack gas, dry



              basis, corrected to standard conditions, g/dscm (g/dscf).



     I      = Percent of isokinetic sampling.

-------
            i-no
                                    x
                                    o
                                   JO
                                    O)
                                   u
                                   ro
                                   0)
                                   0)
                                   l_
                                   D
                                   cn
a: l-

-------
                        i-m
La     = Maximum acceptable leakage rate for either a pretest
         leak check or for a leak check following a component
         change; equal to 0.00057 m /min (0.02 cfm) or 4 percent
         of the average sampling rate, whichever is less.
L.     = Individual leakage rate observed during the leak check
         conducted prior to the "i  " component change (1=1,
         2, 3 ---- n), m /min (cfm).
L      = Leakage rate observed during the post-test leak check,
          o
         m'/m'n (cfm).
m      = Total  amount of parti cul ate matter collected, mg.
M      = Molecular weight of water, 18.0 g/g-mole (18.0 Ib/lb-mole).
 w
m      = Mass of residue of acetone after evaporation, mg.
 a
P.     = Barometric pressure at the sampling site, mm Hg (in. Hg).
P      = Absolute stack gas pressure, mm Hg (in. Hg).
P   .   = Standard absolute pressure, 760 mm Hg (29.92 1n.  Hg).
R      = Ideal  gas constant, 0.06236 mm Hg-m /°K-g-mole (21.85 in.
         Hg-ft3/°R-lb-mole).
T      = Absolute average dry gas meter temperature (see Figure 5-2),
         °
          K
T      = Absolute average stack gas temperature (see Figure 5-2),
         °
          K
T  d   = Standard absolute temperature, 293°K (528°R).
V      = Volume of acetone blank, ml.
 3
V      = Volume of acetone used in wash, ml.
 aW

-------
                          1-112




^lc    = Total volume of liquid collected in impingers and



         silica gel (see Figure 5-3), ml.



V      = Volume of gas sample as measured by dry gas meter,



         dcm (dcf).



V / .  .1= Volume of gas sample measured by the dry gas meter,



         corrected to standard conditions, dscm (dscf).



V /stj\s Volume of water vapor in the gas sample, corrected to



         standard conditions, scm (scf).



v      = Stack gas velocity, calculated by Method 2, Equation 2-9,



         using data obtained from Method  5, m/sec (ft/sec).



W      = Weight of residue in acetone wash, mg.
 3


Y      = Dry gas meter calibration factor.



AH     = Average pressure differential across the orifice meter



         (see Figure 5-2), mm H^O (in. HpO).



p      = Density of acetone, mg/ml (see label on bottle).
 a


p      = Density of water, 0.9982 g/ml (0.002201 Ib/ml).
 W


e      = Total sampling time, min.



e,     = Sampling time interval, from the beginning of a  run until



         the first component change, min.



e.     = Sampling time interval, between  two successive  component



         changes, beginning with the interval between the first



         and second changes, min.



e      = Sampling time interval, from the final (n  ) component



         change until the end of the sampling run, min.

-------
                              1-113
     13.6  = Specific gravity of mercury.
     60    = Sec/min.
     100   = Conversion to percent.
     6.2  Average dry gas meter temperature and average orifice
pressure drop.  See data sheet (Figure 5-2).
     6.3  Dry Gas Volume.  Correct the sample volume measured by the
dry gas meter to standard conditions (20°C, 760 mm Hg or 68°F,
29.92 in. Hg) by using Equation 5-1.
                          Pbar
                                   ^       ,V_Y
Vstd) = VlT

                              p
                               std
                                           1m   	   -.
                                                   m
                                                      Equation 5-1
where:
     KI = 0.3858 °K/mm Hg for metric units
        = 17.64 °R/in. Hg for English units
     Note:  Equation 5-1 can be used as written unless the leakage
rate observed during any of the mandatory leak checks (i.e., the
post-test leak check or leak checks conducted prior to component
changes) exceeds L .  If L  or 1_. exceeds L , Equation 5-1 must be
                  a       p      i          a
modified as follows:
     (a)  Case I.  No component changes made during sampling run.   In
this case, replace V  in Equation 5-1 with the expression:
          Cy» - (LP - La> 6]

-------
                               1-114
     (b)  Case II.  One or more component changes made during the
sampling run.  In this case, replace V  in Equation 5-1  by the
expression:
     r\/  -(L  - L ) 6  -  i  (L   - L )  6  -(L  - L ) 6 1
       m     1    a   1   .  2   i     a   i     p    a   p
and substitute only for those leakage rates (L. or L ) which exceed

     6.4  Volume of water vapor.
                                       /,             Equation 5-2

where:
     1C = 0.001333 m3/ml  for metric units
        = 0.04707 ft3/ml  for English units.
     6.5  Moisture Content.
     Bws  =  V	                       Equation 5-3
      ws     Vm(std)    w(std)
     Note:  In saturated or water droplet-laden  gas  streams,  two
calculations of the moisture content of the stack  gas shall  be made,
one from the impinger analysis (Equation 5-3), and a second  from the
assumption of saturated conditions.   The lower of  the two  values of
B,,c shall be considered correct.   The procedure  for  determining the
 ws
moisture content based upon assumption of saturated  conditions is
given in the Note of Section 1.2  of  Method 4.  For the purposes of this
method, the average stack gas temperature from Figure 5-2  may be used  to
make this determination, provided that the accuracy  of the in-stack
temperature sensor is +_ 1°C (2°F).

-------
                           1-115



     6.6  Acetone Blank Concentration.


                 m
     C      =  V, p,                                     Equation  5-4
      a         a  a



     6.7  Acetone Wash Blank.




     Wa  =  Ca Vaw pa                                    Equation  5-5





     6.8  Total Particulate Weight.   Determine the  total  particulate



catch from the sum of the weights obtained  from containers  1 and 2



less the acetone blank (see Figure 5-3).  Note:   Refer  to Section



4.1.5 to assist in calculation of results involving two or  more



filter assemblies or two or more sampling trains.



     6.9  Particulate Concentration.



     cs  =  (0.001 g/mg)  (%/Vm(std))                    Equation  5-6




     6.10 Conversion Factors:



     From                    Tp_                     Multiply by



     scf                     m3                      0.02832



     g/ft3                   gr/ft3                  15.43



     g/ft3                   lb/ft3                   2.205 x 10"3



     g/ft3                   g/m3                   35.31



     6.11  Isokinetic Variation.



     6.11.1  Calculation From Raw Data.





           100 Ts [K3 V1c + (V, Y/TJ   (Pbar + AH/13.6)]

     1  =  	

                           60 6 v$ PS An




                                                         Equation  5-7

-------
                                1-116
where:
                          •5
     1C = 0.003454 mm Hg-m /ml-°K for metric units
        = 0.002669 in.  Hg-ft3/ml-°R for English units.
     6.11.2  Calculation From Intermediate Values.
     T           Ts Vm(std) Pstd 1QO
        "  Tstd Vs 6 An Ps 60 n-Bws
        = K4  P             s  )-                     Equati°n 5-s
               s  s  n       ws
where:
     K4 = 4.320 for metric units
        = 0.09450 for English units.
     6.12  Acceptable Results.  If 90 percent <_ I <_ 110 percent, the
results are acceptable.  If the results are low in comparison to the
standard  and I is beyond the acceptable range, or, if I is less than
90 percent, the Administrator may opt to accept the results.  Use
Citation 4 to make judgments.  Otherwise, reject the results and repeat
the test.
7.  Bibliography
     1.  Addendum to Specifications for Incinerator Testing at Federal
Facilities.  PHS, NCAPC.  Dec. 6, 1967.
     2.  Martin, Robert M.  Construction Details of Isokinetic Source-
Sampling Equipment.  Environmental Protection Agency.  Research
Triangle Park, N. C.  APTD-0581 .  April, 1971.
     3.  Rom, Jerome J.  Maintenance, Calibration, and Operation
of Isokinetic Source Sampling Equipment.  Environmental Protection
Agency.  Research Triangle Park, N. C.  APTD-0576.  March, 1972.

-------
                              1-117
     4.   Smith, W.  S., R.  T.  Shigehara, and W.  F.  Todd.   A Method
of Interpreting Stack Sampling Data.   Paper Presented at the
63d Annual  Meeting  of the  Air Pollution Control  Association,
St. Louis,  Mo.   June 14-19, 1970.
     5.   Smith, W.  S., et  al.  Stack  Gas Sampling  Improved and
Simplified  With New Equipment.  APCA  Paper No.  67-119.   1967.
     6.   Specifications for Incinerator Testing  at Federal Facilities.
PHS, NCAPC.  1967.
     7.   Shigehara, R.T.  Adjustments in the EPA Nomograph for
Different Pi tot Tube Coefficients  and Dry Molecular Weights.  Stack
Sampling News 2;4-11.  October, 1974.
     8.   Vollaro, R. F.  A Survey  of  Commercially  Available Instrumentation
For the Measurement of Low-Range Gas  Velocities.  U. S.  Environmental
Protection  Agency,  Emission Measurement Branch.   Research Triangle
Park, N. C.  November, 1976 (unpublished paper).
     9.   Annual Book of ASTM Standards.  Part 26.   Gaseous Fuels;
Coal and Coke;  Atmospheric Analysis.   American Society for Testing
and Materials.   Philadelphia, Pa.   1974.  pp. 617-622.

-------
DETERMINATION OF SULFUR DIOXIDE EMISSIONS



        FROM STATIONARY SOURCES





               METHOD 6

-------

-------
                              1-119
            METHOD 6—DETERMINATION OF SULFUR DIOXIDE
                EMISSIONS FROM STATIONARY SOURCES
1.   Principle and Applicability
    1.1.  Principle.   A gas sample is extracted from the sampling
point in the stack.   The sulfuric acid mist (including sulfur trioxide)
and the sulfur dioxide are separated.  The sulfur dioxide fraction
is measured by the barium-thorin titration method.
    1.2  Applicability.  This method is applicable for the determina-
tion of sulfur dioxide emissions from stationary sources.  The
minimum detectable limit of the method has been determined to be
                            3           -73
3.4 milligrams (mg)  of S02/m  (2.12 x 10   Ib/ft ).  Although no upper
limit has been established, tests have shown that concentrations as high
              3
as 80,000 mg/m  of S02 can be collected efficiently in two midget impingers,
each containing 15 milliliters of 3 percent hydrogen peroxide, at a rate
of 1.0 1pm for 20 minutes.  Based on theoretical calculations, the upper
concentration limit in a 20-liter sample is about 93,300 mg/m .
    Possible interferents are free ammonia, water-soluble cations,
and fluorides.  The cations and fluorides are removed by glass wool
filters and an isopropanol bubbler, and hence do not affect the SOp
analysis.  When samples are being taken from a gas stream with high
concentrations of very fine metallic fumes (such as in inlets to
control devices), a high-efficiency glass fiber filter must be used
in place of the glass wool plug (i.e., the one in the probe) to remove
the cation interferents.
    Free ammonia interferes by reacting with S02 to form particulate
sulfite and by reacting with the indicator.  If free ammonia is

-------
                          1-120
     co
     co
     <
     _j
     O
>£
o
<
CO
V
 B§~

 2sg
 Q*
 i£O2
 iuT55
 sg-
 cc5
 0.
            ec
            uu
            I
(O

0>

a
O)
A
                      \

-------
                               1-121
present (this can be determined by knowledge of the process  and
noticing white participate matter in the probe and isopropanol
bubbler), alternative methods, subject to the approval  of the
Administrator, U. S. Environmental Protection Agency,  are required.
2.  Apparatus
     2.1  Sampling.  The sampling train is shown in Figure 6-1,
and component  parts are discussed below.  The tester has the
option of substituting sampling equipment described in Method 8
in place of the midget impinger equipment of Method 6.   However, the
Method 8 train must be modified to include a heated filter between
the probe and isopropanol impinger, and the operation of the sampling
train and sample analysis must be at the flow rates and solution
volumes defined  in Method 8.
    The tester also has the option of determining SCL simultaneously
with particulate matter and moisture determinations by (1) replacing
the water in a Method 5 impinger system with 3 percent peroxide  solution,
or (2) by replacing the Method 5 water impinger system with a Method 8
isopropanol-filter-peroxide system.  The analysis for SCL must be con-
sistent with the procedure in Method 8.
    2.1.1  Probe.   Borosilicate glass, or stainless steel (other
materials of construction may be used, subject to the approval of
the Administrator), approximately 6-mm inside diameter, with a heating
system to prevent  water condensation and a filter (either in-stack or
heated out-stack)  to  remove particulate matter, including sulfuric
acid mist.  A plug of glass wool  is a satisfactory filter.
     2.1.2   Bubbler and  Impingers.  One midget bubbler, with
medium-coarse glass frit  and  borosilicate or quartz glass wool

-------
                             1-122
packed in top (see Figure 6-1)  to prevent sulfuric acid mist
carryover, and three 30-ml midget impingers.   The bubbler and
midget impingers must be connected in series  with leak-free glass
connectors. Silicone grease may be used, if necessary, to prevent
leakage.
     At the option of the tester, a midget impinger may be used
in place of the midget bubbler.
     Other collection absorbers and flow rates may be used, but
are subject to the approval of the Administrator.  Also, collection
efficiency must be shown to be at least 99 percent for each test run
and must be documented in the report.  If the efficiency is found to
be acceptable after a series of three tests,  further documentation is
not required.  To conduct the efficiency test, an extra absorber must
be added and analyzed separately.  This extra absorber must not contain
more than 1 percent of the total SCL.
     2.1.3  Glass Wool.  Borosilicate or quartz.
     2.1.4  Stopcock Grease.  Acetone-insoluble,  heat-stable
silicone grease may be used, if necessary.
     2.1.5  Temperature Gauge.   Dial thermometer, or equivalent, to
measure temperature of gas leaving impinger train to within 1°C (2°F).
     2.1.6  Drying Tube.  Tube packed with 6- to  16-mesh indicating-
type silica gel, or equivalent, to dry the gas sample and to protect
the meter and pump.  If the silica gel has been used previously, dry
at 175°C (350°F) for 2 hours.  New silica gel may be used as received.
Alternatively, other types of desiccants (equivalent or better) may
be used, subject to approval of the Administrator.

-------
                                1-123
     2.1.7  Valve.   Needle valve, to regulate sample gas flow
rate.
     2.1.8  Pump.  Leak-free diaphragm pump, or equivalent, to
pull gas through the train.  Install a small surge tank between the
pump and rate meter to eliminate the pulsation effect of the
diaphragm pump on the rotameter.
     2.1.9  Rate Meter.  Rotameter, or equivalent, capable of
measuring flow rate to within 2 percent of the selected flow rate
of about 1000 cc/min.
     2.1.10  Volume Meter.  Dry gas meter, sufficiently accurate to
measure the sample volume within 2 percent, calibrated at the selected
flow rate and conditions actually encountered during sampling, and
equipped with a temperature gauge (dial thermometer, or equivalent)
capable of measuring temperature to within 3°C (5.4°F).
    2.1.11  Barometer.  Mercury, aneroid, or other barometer
capable of measuring atmospheric pressure to within 2.5 mm Hg
(0.1 in. Hg). In many cases, the barometric reading may be obtained
from a nearby national weather service station, in which case the
station value (which is the absolute barometric pressure) shall
be requested and an adjustment for elevation differences between
the weather station and sampling point shall be applied at a rate
of minus 2.5 mm Hg (0.1 in. Hg) per 30 m (100 ft) elevation
increase or vice versa for elevation decrease.

-------
                              1-124
     2.1.12  Vacuum Gauge and Rotameter.  At least 760 mm Hg (30 in. Hg)
gauge and 0-40 cc/min rotameter, to be used for leak check of the sampling
train.
     2.2  Sample Recovery.
     2.2.1  Wash Bottles.  Polyethylene or glass, 500 ml, two.
     2.2.2  Storage Bottles.  Polyethylene, 100 ml, to store impinger
samples (one per sample).
     2.3  Analysis.
     2.3.1  Pipettes.  Volumetric type, 5-ml, 20-ml (one per sample),
and 25-ml sizes.
     2.3.2  Volumetric Flasks.  100-ml size (one per sample) and 1000-ml
size.
     2.3.3  Burettes. -5- and 50-ml  sizes.
     2.3.4  Erlenmeyer Flasks.  250 mi-size (one for each sample,
blank, and standard).
     2.3.5  Dropping Bottle.  125-ml size, to add indicator.
     2.3.6  Graduated cylinder.   100-ml size.
     2.3.7  Spectrophotometer.  To measure absorbance at 352 nanometers.
3.  Reagents
     Unless otherwise indicated, all reagents  must conform to the
specifications established by the Committee on Analytical  Reagents  of
the American Chemical Society.  Where such specifications  are not
available, use the best available grade.
     3.1  Sampling.
     3.1.1  Water.  Deionized, distilled to conform to ASTM
specification D1193-74, Type 3.   At  the option of the analyst,

-------
                                1-125
the KMnO.  test for oxidizable organic matter may be omitted when high
concentrations of organic matter are not expected to be present.
     3.1.2 Isopropanol, 80 percent.  Mix 80 ml of isopropanol with 20
ml of deionized, distilled water.  Check each lot of isopropanol for
peroxide impurities as follows:  shake 10 ml of isopropanol with 10 ml
of freshly prepared 10 percent potassium iodide solution.  Prepare a
blank by similarly treating 10 ml of distilled water.  After 1 minute,
read the absorbance at 352 nanometers on a spectrophotometer (Note:  Use
a 1-cm path length).  If absorbance exceeds 0.1, reject
alcohol for use.
     Peroxides may be removed from isopropanol by redistilling or by
passage through a column of activated alumina; however, reagent grade
isopropanol with suitably low peroxide levels may be obtained from
commercial sources.-  Rejection of contaminated lots may, therefore, be a
more efficient procedure.
     3.1.3 Hydrogen Peroxide, 3 Percent.  Dilute 30 percent hydrogen
peroxide 1:9 (v/v) with deionized, distilled water (30 ml is needed per
sample).   Prepare fresh daily.
     3.1.4  Potassium Iodide Solution, 10 Percent.     Dissolve 10.0
grams KI in deionized, distilled water and dilute to 100 ml.   Prepare
when needed.
     3.2   Sample Recovery.
     3.2.1  Water.  Deionized, distilled, as in 3.1.1.
     3.2.2  Isopropanol, 80 Percent.   Mix 80 ml of isopropanol with 20
ml of deionized, distilled water.

-------
                              1-126
     3.3  Analysis.
     3.3.1  Water.   Deionized,  distilled,  as  in  3.1.1.
     3.3.2  Isopropanol,  100 percent.
     3.3.3  Thorin  Indicator.   l-(o-arsonophenylazo)-2-naphthol-3,
6-disulfonic acid,  disodium salt,  or equivalent.   Dissolve  0.20  g
in 100 ml  of deionized,  distilled  water.
     3.3.4  Barium  Perchlorate  Solution,  0.0100  N.   Dissolve  1.95 g  of
barium perchlorate  trihydrate [Ba(C104)2'3H20] in 200 ml  distilled water
and dilute to 1  liter with isopropanol.   Alternatively,  1.22  g of
[BaCl2'2H20] may be used instead of  the  perchlorate.  Standardize as in
Section 5.5.
     3.3.5  Sulfuric Acid Standard,  0.0100 N.  Purchase  or  standardize
to +0.0002 N against 0.0100 N NaOH which  has  previously  been
standardized against potassium acid  phthalate  (primary standard
grade).
4.  Procedure.
     4.1  Sampling.
     4.1.1  Preparation of collection train.   Measure 15 ml of 80
percent isopropanol into the midget  bubbler and  15 ml of 3  percent
hydrogen peroxide into each of the first  two  midget impingers.
Leave the final  midget impinger dry.  Assemble the train as shown
in Figure 6-1.  Adjust probe heater  to a  temperature sufficient  to
prevent water condensation.  Place crushed ice and water around  the
impingers.
     4.1.2  Leak-check procedure.   A leak check  prior to the
sampling run is optional; however, a leak check  after the sampling
run is mandatory.  The leak-check  procedure is as follows:

-------
                                 1-127
     Temporarily attach a suitable (e.g., 0-40 cc/min) rotameter to the
outlet of the dry gas meter and place a vacuum gauge at or near the
probe inlet.  Plug the probe inlet, pull a vacuum of at least 250 mm Hg
(10 in. Hg), and note the flow rate as indicated by the rotameter.  A
leakage rate not in excess of 2 percent of the average sampling rate is
acceptable.  Note:  Carefully release the probe inlet plug before
turning off the pump.
     It is suggested (not mandatory) that the pump be leak-checked
separately, either prior to or after the sampling run.  If done prior-to
the sampling run, the pump leak-check shall precede the leak check of
the sampling train described immediately above; if done after the
sampling run, the pump leak-check shall follow the train leak-check.   To
leak check the pump, proceed as follows:  Disconnect the drying tube
from the probe-impinger assembly.  Place a vacuum gauge at the Inlet to
either the drying tube or the pump, pull a vacuum of 250 mm (10 in.)  Hg,
plug or pinch off the outlet of the flow meter and then turn off the
pump.  The vacuum should remain stable for at least 30 seconds.
     Other leak-check procedures may be used, subject to the approval of
the Administrator, U.S. Environmental Protection Agency.

-------
                                   1-128
     4.1.3  Sample Collection.  Record the initial  dry gas meter reading
and barometric pressure.  To begin sampling, position the tip of the
probe at the sampling point, connect the probe to the bubbler, and start
the pump.  Adjust the sample flow to a constant rate of approximately
1.0 liter/min as indicated by the rotameter.  Maintain this constant
rate (_+ 10 percent) during the entire sampling run.   Take readings (dry
gas meter, temperatures at dry gas meter and at impinger outlet and rate
meter) at least every 5 minutes.  Add more ice during the run to keep
the temperature of the gases leaving the last impinger at 20°C (68°F) or
less.  At the conclusion of each run, turn off the  pump, remove probe
from the stack, and record the final readings.  Conduct a leak check as
in Section 4.1.2.  (This' leak check is mandatory.)   If a leak is found,
void the test run or use procedures acceptable to the Administrator to
adjust the sample volume for leakage.  Drain the ice bath, and purge the
remaining part of the train by drawing clean ambient air through the
system for 15 minutes at the sampling rate.
     Clean ambient air can be provided by passing air through a charcoal
filter or through an extra midget impinger with 15  ml of 3 percent HpO^.
The tester may opt to simply use ambient air, without purification.

-------
                                    1-129
     4.2  Sample Recovery.   Disconnect the impingers after purging.
Discard the contents of the midget bubbler.   Pour the contents of
the midget impingers into a leak-free polyethylene bottle for
shipment.  Rinse the three midget impingers  and the connecting
tubes with deionized, distilled water, and add the washings to
the same storage container.  Mark the fluid  level.  Seal  and
identify the sample container.
     4.3  Sample Analysis.  Note level of liquid in container,
and confirm whether any sample was lost during shipment;  note
this on analytical data sheet.  If a noticeable amount of leakage has
occurred, either void the sample or use methods, subject  to the approval
of the Administrator, to correct the final results.
     Transfer the contents of the storage container to a  100-ml
volumetric flask and dilute to exactly 100 ml with deionized, distilled
water.  Pipette a 20-ml aliquot of this solution into a 250-ml
Erlenmeyer flask, add 80 ml of 100 percent isopropanol and two to four
drops of thorin indicator, and titrate to a pink endpoint using
0.0100 N barium perchlorate.  Repeat and average the titration
volumes.  Run a blank with each series of samples.  Replicate
titrations must agree within 1 percent or 0.2 ml, whichever is larger.
      (Note:  Protect the 0.0100 N barium perchlorate solution from
evaporation at all times.)

-------
                              1-130
5.  Calibration
     5.1  Metering System.
     5.1.1  Initial Calibration.  Before its initial use in the field,
first leak check the metering system (drying tube, needle valve, pump,
rotameter, and dry gas meter) as follows:  place a vacuum gauge at
the inlet to the drying tube and pull a vacuum of 250 mm (10 in.) Hg;
plug or pinch off the outlet of the flow meter, and then turn off the
pump.  The vacuum shall remain stable for at least 30 seconds.   Care-
fully release the vacuum gauge before releasing the flow meter end.
     Next, calibrate the metering system (at the sampling flow rate
specified by the method) as follows:  connect an appropriately sized
wet test meter (e.g., 1 liter per revolution) to the inlet of the
drying tube.  Make three independent calibration runs, using at least
five revolutions of the dry gas meter per run.   Calculate the calibra-
tion factor, Y (wet test meter calibration volume divided by the dry
gas meter volume, both volumes adjusted to the  same reference temperature  and
pressure), for each run, and average the results.   If any Y value
deviates by more than 2 percent from the average, the metering  system
is unacceptable for use.  Otherwise, use the average as  the calibration
factor for subsequent test runs.
     5.1.2  Post-Test Calibration Check.  After each field test series,
conduct a calibration check as in Section 5.1.1  above, except for the
following variations: (a) the leak check is not to be conducted, (b)  three,
or more revolutions of the dry gas meter may be used, and (c) only
two Independent runs need be made.  If the calibration factor does

-------
                             1-131
not deviate by more than 5 percent  from the initial  calibration
factor (determined in Section 5.1.1),  then the dry gas meter
volumes obtained during the test series are acceptable.  If the
calibration factor deviates by more than 5 percent,  recalibrate
the metering system as in Section 5.1.1, and for the calculations,
use the calibration factor (initial or recalibration) that yields the
lower gas volume for each test run.
     5.2  Thermometers.  Calibrate  against mercury-in-glass thermometers.
     5.3  Rotameter.  The rotameter need not be calibrated, but should
be cleaned and maintained according to the manufacturer's instruction.
     5.4  Barometer.  Calibrate against a mercury barometer.
     5.5  Barium Perchlorate Solution.  Standardize  the barium perch-
lorate solution against 25 ml of standard sulfuric acid to which 100 ml
of 100 percent isopropanol has been added.
6.  Calculations
    Carry out calculations, retaining  at least one extra decimal figure
beyond that of the acquired data.  Round off figures after final calculation.
    6.1  Nomenclature.
    CSQ     = Concentration of sulfur  dioxide, dry basis corrected  to
               standard conditions, mg/dscm (Ib/dscf).
    N       = Normality of barium perchlorate titrant, mi Hi equivalents/ml.
    P.       = Barometric pressure at the exit orifice of the dry gas
     oar
              meter, mm Hg (in. Hg).
    Pstd    = standard absolute pressure, 760 mm Hg (29.92 in. Hg).
    Tm      = Avera9e dr%y 9as meter absolute temperature, °K (°R).
    Tstd    = standard absolute temperature, 293° K (528° R).

-------
                              1-132




    V,      = Volume of sample aliquot titrated, ml.
     a                                          •»


    V       = Dry gas volume as measured by the dry gas meter, dcm



              (dcf).



    V /  .,\ = Dry gas volume measured by the dry gas  meter,



              corrected to standard conditions, dscm  (dscf).



    V  T     = Total  volume of solution in which the sulfur dioxide
     soln


              sample is contained, 100 ml.



    V      = Volume of barium perch!orate titrant used for the



             sample, ml (average of replicate titrations).



    V^.     = Volume of barium perchlorate titrant used for the
     tb


             blank, ml.



    Y      = Dry gas -meter calibration factor.



    32.03  = Equivalent weight of sulfur dioxide.



    6.2  Dry sample gas volume, corrected to standard conditions.




                                                     V  P
                                               „ „    m  bar
                                                     Equation 6-1



where:



     K.J  = 0.3858 °K/mm Hg for metric units.



         = 17.64 °R/in. Hg for English units.



     6.3  Sulfur dioxide concentration.



                                 (/Vsoln

                     (V.  - V,, )N\~T

     Ccn    =   K0   —	rr^	-	^-             Equation 6-2

      bU2        2         Vm(std)



where:



     Kp  =  32.03 mg/meq. for metric units.




         = 7.061 x 10"  Ib/meq. for English  units.

-------
                                1-133
 7.  Bibliography
     1.  Atmospheric Emissions from Sulfuric Acid Manufacturing
Processes.  U.S. DHEW, PHS, Division of Air Pollution.  Public
Health Service Publication No. 999-AP-13.  Cincinnati, Ohio.  1965.
     2.  Corbett, P.P.  The Determination of SO^ and S(L in Flue
Gases.  Journal  of the Institute of Fuel.  2£:237-243, 1961.
     3.  Matty,  R. E. and E. K. Diehl.  Measuring Flue-Gas S02 and
SO..  Power.  l_01_:94-97.  November 1957.
     4.  Patton, W. F. and J.  A. Brink, Jr.  New Equipment and
Techniques for Sampling Chemical Process Gases.  J. Air Pollution
Control Association.  1_3:162.   1963.
     5.  Rom, J. J.  Maintenance, Calibration, and Operation of
Isokinetic Source-Sampling Equipment.   Office of Air Programs,
Environmental Protection Agency.  Research Triangle Park, N. C.
APTD-0576.  March 1972.
     6.  Hamil,  H. F. and D. E. Camann.  Collaborative Study of
Method for the Determination of Sulfur Dioxide Emissions From
Stationary Sources (Fossil-Fuel Fired Steam Generators).
Environmental Protection Agency, Research Triangle Park, N. C.
EPA-650/4-74-024.  December 1973.
     7.  Annual  Book of ASTM Standards.  Part 31; Water, Atmospheric
Analysis.  American Society for Testing and Materials.  Philadelphia,
PA.  1974.  pp.  40-42.
     8.  Knoll,  J. E. and M. R. Midgett.  The Application of EPA
Method 6 to High Sulfur Dioxide Concentrations.  Environmental Protection
Agency.  Research Triangle Park, N. C.  EPA-600/4-76-038.  July 1976.

-------
DETERMINATION OF NITROGEN OXIDE EMISSIONS



        FROM STATIONARY SOURCES





                METHOD?

-------
                               1-135


              METHOD 7—DETERMINATION OF NITROGEN OXIDE

                  EMISSIONS FROM STATIONARY SOURCES
1.   Principle and Applicability



     1.1  Principle.  A grab sample is collected in an evacuated



flask containing a dilute sulfuric acid-hydrogen peroxide absorbing



solution, and the nitrogen oxides, except nitrous oxide, are



measured colorimetrically using the phenoldisulfonic acid (PDS)



procedure.



     1.2  Applicability.  This method is applicable to the measure-



ment of nitrogen oxides emitted from stationary sources.  The range



of the method has been determined to be 2 to 400 milligrams NO  (as
                                                              A


NO,,) per dry standard cubic meter, without having to dilute the sample.



2.   Apparatus



     2.1  Sampling (see Figure 7-1).  Other grab sampling systems or



equipment, capable of measuring sample volume to within +2.0 percent



and collecting a sufficient sample volume to allow analytical reproduci-



bility to within +5 percent, will be considered acceptable alternatives,



subject to approval of the Administrator, U. S. Environmental Protection



Agency.  The following equipment is used in sampling:



     2.1.1  Probe.  Borosilicate glass tubing, sufficiently heated to



prevent water condensation and equipped with an in-stack or out-stack



filter to remove particulate matter (a plug of glass wool is satisfactory



for this purpose).  Stainless steel or Teflon  tubing may also be used



for the probe.  Heating is not necessary if the probe remains dry



during the purging period.
 Mention of trade names or specific products does not constitute

 endorsement by the Environmental Protection Agency.

-------
1-136
                                    o
                                    LU
                                             (0
                                             re
                                             re
                                             c

                                             5
                                             Q.


                                             re
                                            CO
                                             0)

                                             3

                                             O>

-------
                               1-137
     2.1.2  Collection Flask.   Two-liter borosilicate, round bottom
flask, with short neck and 24/40  standard taper opening,  protected
against implosion or breakage.
     2.1.3  Flask Valve.   T-bore  stopcock connected to a  24/40
standard taper joint.
     2.1.4  Temperature Gauge.  Dial-type thermometer, or other
temperature gauge, capable of measuring 1°C (2°F)  intervals  from -5
to 50°C (25 to 125°F).
     2.1.5  Vacuum Line.   Tubing  capable of withstanding  a vacuum
of 75 mm Hg (3 in. Hg) absolute pressure, with "T" connection and
T-bore stopcock.
     2.1.6  Vacuum Gauge.   U-tube manometer, 1 meter (36  in.), with
1-mm (0.1-in.) divisions,  or other gauge capable of measuring pressure
to within +2.5 mm Hg (0.10 in.  Hg).
     2.1.7  Pump.  Capable of evacuating the collection flask to a
pressure equal to or less  than 75 mm Hg (3 in. Hg) absolute.
     2.1.8  Squeeze Bulb.   One-way.
     2.1.9  Volumetric Pipette.  25ml.
     2.1.10  Stopcock and  Ground  Joint Grease.  A high-vacuum, high-
temperature chlorofluorocarbon grease is required.  Halocarbon 25-5S
has been found to be effective.
     2.1.11  Barometer.  Mercury, aneroid, or other barometer capable
of measuring atmospheric pressure to within 2.5 mm Hg (0.1 in. Hg).
In many cases, the barometric reading may be obtained from a nearby
national weather service station, in which case the station  value

-------
                              1-138
(which is the absolute barometric pressure)  shall  be requested and
an adjustment for elevation differences between the weather station
and sampling point shall  be applied at a rate of minus 2;5 mm Hg
(0.1 in. Hg) per 30 m (100 ft) elevation increase, or vice versa
for elevation decrease.
     2.2  Sample Recovery.  The following equipment is required for
sample recovery:
     2.2.1  Graduated Cylinder.   50 ml with  1-ml divisions.
     2.2.2  Storage Containers.   Leak-free polyethylene bottles.
     2.2.3  Wash Bottle.   Polyethylene or glass.
     2.2.4  Glass Stirring Rod.
     2.2.5  Test Paper for Indicating pH.  To cover the pH range of
7 to 14.
     2.3  Analysis.  For the analysis, the following equipment is needed:
     2.3.1  Volumetric Pipettes.  Two 1 ml,  two 2 ml, one 3 ml, one
4 ml, two 10 ml, and one 25 ml for each sample and standard.
     2.3.2  Porcelain Evaporating Dishes.  175- to 250-ml capacity
with lip for pouring, one for each sample and each standard.  The
Coors No. 45006 (shallow-form, 195 ml) has been found to be satisfactory.
Alternatively, polymethyl pentene beakers (Nalge No. 1203, 150 ml), or
glass beakers (150 ml) may be used.  When glass beakers are used, etching
of the beakers may cause solid matter to be present in the analytical
step; the solids should be removed by filtration (see Section 4.3).
     2.3.3  Steam Bath.  Low-temperature ovens or thermostatically
controlled hot plates kept below 70°C (160°F) are acceptable alternatives.

-------
                               1-139
     2.3.4  Dropping Pipette or Dropper.  Three required.
     2.3.5  Polyethylene Policeman.  One for each sample and each
standard.
     2.3.6  Graduated Cylinder.  100 ml with 1-ml divisions.
     2.3.7  Volumetric Flasks.  50 ml (one for each sample and each
standard), 100 ml (one for each sample and each standard, and one for
the working standard KNO-j solution), and 1000 ml (one).
     2.3.8  Spectrophotometer.  To measure absorbance at 410 nm.
     2.3.9  Graduated Pipette.  10 ml with 0.1-ml divisions.
     2.3.10  Test Paper for Indicating pH.  To cover the pH range of 7
to 14.
     2.3.11  Analytical Balance.  To measure to within 0.1 mg.
3.  Reagents
     Unless otherwise indicated, it is intended that all reagents
conform to the specifications established by the Committee on Analytical
Reagents of the American Chemical Society, where such specifications are
available; otherwise, use the best available grade.
     3.1  Sampling.  To prepare the absorbing solution, cautiously add
2.8 ml concentrated H^SO. to 1 liter of deionized,  distilled water.  Mix
well and add 6 ml of 3 percent hydrogen peroxide, freshly prepared from
30 percent hydrogen peroxide solution.  The absorbing solution should be
used within 1 week of its preparation.  Do not expose to extreme heat or
direct sunlight.
     3.2  Sample Recovery.  Two reagents are required for sample
recovery:

-------
                             1-140

     3.2.1   Sodium Hydroxide (1  N_).   Dissolve  40 g  NaOH  in  deionized,

distilled water and dilute to 1  liter.

     3.2.2  Water.  Deionized,  distilled to conform to ASTM specifi-

cation Dll93-74, Type 3.   At the option of the analyst,  the KMNO.  test

for oxidizable organic matter may be omitted when high concentrations

of organic matter are not expected to be present.

     3.3  Analysis.  For  the analysis,  the following reagents  are

required:

     3.3.1   Fuming Sulfuric Acid.  15 to 18 percent by weight  free

sulfur trioxide.  HANDLE  WITH CAUTION.

     3.3.2  Phenol.  White solid.

     3.3.3  Sulfuric Acid.  Concentrated, 95 percent minimum assay.

HANDLE WITH CAUTION.

     3.3.4  Potassium Nitrate.   Dried at 105 to 110°C (220  to  230°F)

for a minimum of 2 hours  just prior to preparation  of standard solution.

     3.3.5  Standard KN03 Solution.   Dissolve  exactly 2.198 g  of dried

potassium nitrate (KNO-)  in deionized,  distilled water and  dilute

to 1 liter with deionized, distilled water in  a 1000-ml  volumetric

flask.
    •
     3.3.6  Working Standard KNO- Solution. Dilute 10 ml of the

standard solution to 100  ml with deionized distilled water. One

mi Hi liter of the working standard solution is equivalent to 100 pg

nitrogen dioxide (NOp).

     3.3.7  Water.  Deionized,  distilled as in Section 3.2.2.

-------
                                1-141
     3.3.8  Phenoldisulfonic Acid Solution.  Dissolve 25 g of pure
white phenol in 150 ml concentrated sulfuric acid on a steam bath.
Cool, add 75 ml fuming sulfuric acid, and heat at 100°C (212°F) for
2 hours.  Store in a dark, stoppered bottle.
     3.3.9  Concentrated Ammonium Hydroxide.  Reagent grade.
4.  Procedures
     4.1  Sampling.
     4.1.1  Pipette 25 ml of absorbing solution into a sample flask,
retaining a sufficient quantity for use in preparing the calibration
standards.  Insert the flask valve stopper into the flask with the
valve in the "purge" position.  Assemble the sampling train as shown
in Figure 7-1 and place the probe at the sampling point.   Make sure
that all fittings are tight and leak-free, and that all  ground glass
joints have been proper-ly-greased with a high-vacuum, high-temperature
chlorofluorocarbon-based stopcock grease.   Turn the flask valve and the
pump valve to their "evacuate" positions.   Evacuate the  flask to
75 mm Hg (3 in. Hg) absolute pressure, or less.  Evacuation to a
pressure approaching the vapor pressure of water at the  existing
temperature is desirable.  Turn the pump valve to its "vent" position
and turn off the pump.   Check for leakage by observing the manometer
for any pressure fluctuation.  (Any variation greater than 10 mm Hg
(0.4 in. Hg) over a period of 1 minute is  not acceptable,  and the
flask is not to be used until the leakage  problem is corrected.
Pressure in the flask is not to exceed 75  mm Hg (3 in. Hg)  absolute
at the time sampling is commenced.)   Record the volume of the flask
and valve (Vf), the flask temperature (T.), and the barometric pres-
sure.   Turn the flask valve counterclockwise to its "purge" position

-------
                             1-142
and do the same with the pump valve.   Purge the probe and the vacuum
tube using the squeeze bulb.   If condensation occurs in the probe
and the flask valve area, heat the probe and purge until  the con-
densation disappears.  Next,  turn the pump valve to its "vent" position.
Turn the flask valve clockwise to its "evacuate" position and record
the difference in the mercury levels  in the manometer.  The absolute
internal pressure in the flask (P.) is equal to the barometric pres-
sure less the manometer reading.  Immediately turn the flask valve to
the "sample" position and permit the  gas to enter the flask until
pressures in the flask and sample line (i.e., duct, stack) are
equal.  This will usually require about 15 seconds; a longer period
indicates a "plug" in the probe, which must be corrected before
sampling is continued.  After collecting the sample, turn the flask
valve to its "purge" position and disconnect the flask from the
sampling train.  Shake the flask for  at least 5 minutes.
     4.1.2  If the gas being sampled  contains insufficient oxygen
for the conversion of NO to N0? (e.g., an applicable subpart of the
standard may require taking a sample  of a calibration gas mixture of NO
in Np), then oxygen shall be introduced into the flask to permit this
conversion.  Oxygen may be introduced into the flask by one of
three methods:   (1) Before evacuating the sampling flask, flush
with pure cylinder oxygen, then evacuate flask to 75 mm Hg (3 in.
Hg) absolute pressure or less; or  (2) inject oxygen into the flask
after sampling; or (3) terminate sampling with a minimum of 50 mm Hg
(2 in. Hg) vacuum remaining in the flask, record this final pressure,
and then vent the flask to the atmosphere until the flask pressure is
almost equal to atmospheric pressure.

-------
                               1-143
     4.2  Sample Recovery.  Let the flask set for a minimum of 16 hours
and then shake the contents for 2 minutes.  Connect the flask to a mercury
filled U-tube manometer.  Open the valve from the flask to the
manometer and record the flask temperature (T^), the barometric
pressure, and the difference between the mercury levels in the
manometer.  The absolute internal pressure in the flask (PJ is
the barometric pressure less the manometer reading.  Transfer the
contents of the flask to a leak-free polyethylene bottle.   Rinse
the flask twice with 5-ml portions of deionized, distilled water
and add the rinse water to the bottle.  Adjust the pH to between 9 and
12 by adding sodium hydroxide (1  Nj, dropwise (about 25 to 35 drops).
Check the pH by dipping a stirring rod into the solution and then
touching the rod to the pH test paper.  Remove as little material as
possible during this step.  Mark the height of the liquid level so
that the container can be checked for leakage after transport.  Label
the container to clearly identify its contents.  Seal the container
for shipping.
     4.3  Analysis.  Note the level of the liquid in container and con-
firm whether or not any sample was lost during shipment; note this on
the analytical data sheet.  If a noticeable amount of leakage has
occurred, either void the sample or use methods, subject to the approval
of the Administrator, to correct the final results.  Immediately prior
to analysis, transfer the contents of the shipping container to a 50-ml
volumetric flask, and rinse the container twice with 5-ml  portions of
deionized, distilled water.  Add the rinse water to the flask and dilute
to the mark with deionized, distilled water; mix thoroughly.  Pipette

-------
                              1-144
a 25-ml  aliquot into the porcelain evaporating dish.   Return  any
unused portion of the sample to the polyethylene storage bottle.
Evaporate the 25-ml  aliquot to dryness on a steam bath and allow to
cool.  Add 2 ml phenoldisulfonic acid solution to the dried residue and
triturate thoroughly with a polyethylene policeman.   Make sure the solu-
tion contacts all the residue.  Add 1 ml deionized,  distilled water and
four drops of concentrated sulfuric acid.  Heat the  solution  on a
steam bath for 3 minutes with occasional stirring.   Allow the solution
to cool, add 20 ml  deionized, distilled water, mix well  by stirring,
and add concentrated ammonium hydroxide, dropwise, with constant stirring,
until the pH is 10 (as determined by pH paper).  If  the sample contains
solids, these must be removed by filtration (centrifugation is an
acceptable alternative, subject to the approval of the Administrator),
as follows:  filter through Whatman No. 41  filter paper into  a 100-ml
volumetric flask; rinse the evaporating dish with three 5-ml  portions
of deionized, distilled water; filter these three rinses.  Wash the
filter with at least three 15-ml portions of deionized, distilled water.
Add the filter washings to the contents of the volumetric flask and
dilute to the mark with deionized, distilled water.   If solids are
absent, the solution can be transferred directly to  the 100-ml volumetric
flask and diluted to the mark with deionized, distilled water.  Mix
the contents of the flask thoroughly, and measure the absorbance at
the optimum wavelength used for the standards (Section 5.2.1), using the
blank solution as a zero reference.  Dilute the sample and the blank
with equal volumes of deionized, distilled water if  the absorbance
exceeds A., the absorbance of the 400 ug N02 standard (see Section 5.2.2).

-------
                               1-145




5.  Calibration



     5.1  Flask Volume.  The volume of the collection flask-flask valve



combination must be known prior to sampling.  Assemble the flask and



flask valve and fill with water, to the stopcock.  Measure the volume of



water to +10 ml.  Record this volume on the flask.



     5.2  Spectrophotometer Calibration.



     5.2.1  Optimum Wavelength Determination.  Calibrate the wavelength



scale of the spectrophotometer every 6 months.  The calibration may be



accomplished by using an energy source with an intense line emission



such as a mercury lamp, or by using a series of glass filters spanning



the measuring range of the spectrophotometer.  Calibration materials are



available commercially and from the National Bureau of Standards.



Specific details on the use of such materials should be supplied by the



vendor; general information about calibration techniques can be obtained



from general reference books on analytical chemistry.  The wavelength



scale of the spectrophotometer must read correctly within +5 nm at all



calibration points; otherwise, the spectrophotometer shall be repaired



and recalibrated.   Once the wavelength scale of the spectrophotometer is



in proper calibration, use 410 nm as the optimum wavelength for the



measurement of the absorbance of the standards and samples.

-------
                                1-146
     Alternatively, a scanning procedure may be employed to determine
 the proper measuring wavelength.  If the instrument is a double-beam
 spectrophotometer, scan the spectrum between 400 and 415 rim using a 200 yg
 N02 standard solution in the sample cell and a blank solution in the
 reference cell.  If a peak does not occur, the spectrophotometer is
 probably malfunctioning and should be repaired.  When a peak is obtained
within the 400 to 415 nm range, the wavelength at which this peak occurs
 shall be the optimum wavelength for the measurement of absorbance of
 both the standards and the samples.   For a single-beam spectrophotometer,
 follow the scanning procedure described above, except that the blank and
 standard solutions shall be scanned separately.  The optimum wavelength
 shall be the wavelength at which the maximum difference in absorbance
between the standard and -the blank occurs.
     5.2.2  Determination-of Spectrophotometer Calibration Factor K .
Add 0.0 ml, 2.0 ml, 4.0 ml, 6.0 ml,  and 8.0 ml of the KNOj working
standard solution (1  ml = 100 yg IW^) to a series of five 50-ml volumetric
flasks.  To each flask, add 25 ml  of absorbing solution,
10 ml deionized, distilled water,  and sodium hydroxide (IN) dropwise
until the pH is between 9 and 12 (about 25 to 35 drops each).   Dilute
to the mark with deionized, distilled water.   Mix thoroughly and
pipette a 25-ml aliquot of each solution into a separate  porcelain
evaporating dish.  Beginning with  the evaporation step, follow the
analysis procedure of Section 4.3, until the solution has been trans-
ferred to the 100-ml  volumetric flask and diluted to the  mark.   Measure
the absorbance of each solution, at  the optimum wavelength, as
determined in Section 5.2.1.  This calibration procedure  must be repeated

-------
                                1-147



on each day that samples are analyzed.  Calculate the spectrophotometer



calibration factor as follows:



                  A  + 2A, + 3A, + 4A.

         K  = 100 -U - ^5 - z* - \               Equation 7-1

          c                           *
where:
     K  = Calibration factor
     A, = Absorbance of the 100-yg NOp standard



     Ap = Absorbance of the 200-yg NO,, standard



     A^ = Absorbance of the 300-yg N09 standard
      O                              w


     A^ = Absorbance of the 400-yg NO^ standard
     5.3  Barometer.  Calibrate against a mercury barometer.



     5.4  Temperature Gauge.  Calibrate dial thermometers against



mercury-in-glass thermometers.



     5.5  Vacuum Gauge.  Calibrate mechanical gauges, if used,



against a mercury manometer such as that specified in 2.1.6.



     5.6  Analytical Balance.  Calibrate against standard weights.



6.  Calculations



     Carry out the calculations, retaining at least one extra



decimal figure beyond that of the acquired data.  Round off



figures after final calculations.



     6.1  Nomenclature.



     A  = Absorbance of sample.



     C  = Concentration of NO  as N09, dry basis, corrected to
                             A      C


          standard conditions, mg/dscm (Ib/dscf).

-------
                                 1-148
     F   =  Dilution  factor  (i.e.,  25/5,  25/10, etc., required only



          if  sample dilution was  needed to reduce the absorbance



          into  the  range of calibration).



     K   =  Spectrophotometer calibration factor.



     m   =  Mass  of NO  as NC9 in gas  sample, yg.
                    A      C,


     P. =  Final  absolute pressure of flask, mm Hg (in. Hg).



     P. =  Initial absolute pressure  of  flask, mm Hg (in. Hg).



     Pstd  =  Standard absolute  pressure, 760 mm Hg (29.92 in. Hg).



     Tf =  Final  absolute temperature of flask, °K (°R).



     T. =  Initial absolute temperature  of flask, °K (°R).



     T   d  =  Standard absolute  temperature, 293°K (528°R).



     V     =  Sample  volume  at standard conditions (dry basis), ml.



     Vf =  Volume of flask  and  valve, ml.



     V   =  Volume of absorbing  solution, 25 ml.
      a


     2   =  50/25, the aliquot factor.   (If other than a 25-ml



          aliquot was  used for analysis, the  corresponding  factor



          must  be substituted).



     6.2  Sample volume, dry basis,  corrected to standard



conditions.
 sc
Tstd
Pstd

-------
                               1-149
where:
                    °K
     KI  =  0.3858 mff) H   for metric units
                    OR
         = 17.64  -r  K
                  .    j.   for English units
     6.3  Total yg NO^ per sample.
          m = 2«cAF                                    Equation 7-3
     Note:  If other than a 25-ml  aliquot is used for analysis, the
factor 2 must be replaced by a corresponding factor.
     6.4  Sample concentration, dry basis, corrected to standard
conditions.
     C  =  K2 T~                                     Equation 7-4
               sc
where:
                     3
     KO  =  10   ^%r  for metric units
      2          ug/ml
         = 6.243 x 10"5        for English units
7.  Bibliography
     1.  Standard Methods of Chemical Analysis.  6th ed. New York,
D. Van Nostrand Co., Inc.  1962.  Vol. 1, p. 329-330.
     2.  Standard Method of Test for Oxides of Nitrogen in Gaseous
Combustion Products (Phenoldisulfonic Acid Procedure).  In:  1968 Book
of ASTM Standards, Part 26.  Philadelphia, Pa.  1968.  ASTM Designation
D-1608-60, p. 725-729.
     3.  Jacob, M. B.   The Chemical Analysis of Air Pollutants.
New York.  Interscience Publishers, Inc.  1960.  Vol. 10, p. 351-356.

-------
                                1-150
     4.   Beatty, R.  L., L.  B.  Berger, and H.  H.  Schrenk.   Determination
of Oxides of Nitrogen by the Phenoldisulfonic Acid Method.  Bureau of
Mines, U. S. Dept. of Interior.   R.  I. 3687.   February .1943.
     5.   Hamil, H. F. and D. E.  Camann.  Collaborative Study of Method
for the Determination of Nitrogen Oxide Emissions from Stationary
Sources (Fossil Fuel-Fired  Steam Generators).  Southwest Research
Institute report for Environmental  Protection Agency.  Research Triangle
Park, N. C.  October 5, 1973.
     6.   Hamil, H. F. and R. E.  Thomas.  Collaborative Study of Method
for the Determination of Nitrogen Oxide Emissions from Stationary
Sources (Nitric Acid Plants).   Southwest Research Institute report for
Environmental Protection Agency.  Research Triangle Park, N. C.
May 8, 1974.

-------
DETERMINATION OF SULFURIC ACID MIST AND SULFUR DIOXIDE



         EMISSIONS FROM STATIONARY SOURCES





                     METHODS

-------
                                       1-151
          METHOD 8—DETERMINATION OF SULFURIC ACID MIST AND
           SULFUR DIOXIDE EMISSIONS FROM STATIONARY SOURCES

1.  Principle and Applicability
     1.1  Principle.  A gas sample is extracted isokinetically from the
stack.  The sulfuric acid mist (including sulfur trioxide) and the
sulfur dioxide are separated, and both fractions are measured  separately
by the barium-thorin titration method.
     1.2  Applicability.  This method is applicable for the  determi-
nation of sulfuric acid mist (including sulfur trioxide, and in the
absence of other particulate matter) and sulfur dioxide emissions  from
stationary sources.  Collaborative tests have shown that the minimum
detectable limits of the method are 0.05 milligrams/cubic meter (0.03
    -7                  •                                  3           -7
x 10"  pounds/cubic foot) for sulfur trioxide and 1.2 mg/m (0.74 x 10
     2
Ib/ft ) for sulfur .dioxide.  No upper limits have been established.
Based on theoretical calculations for 200 milliliters of 3 percent
hydrogen peroxide solution, the upper concentration limit for  sulfur
dioxide in a 1.0 m  (35.3 ft ) gas sample is about 12,500 mg/m3 (7.7 x
10"  Ib/ft ).  The upper limit can be extended by increasing the
quantity of peroxide solution in the impingers.
     Possible interfering agents of this method are fluorides, free
ammonia, and dimethyl  aniline.  If any of these interfering  agents are
present (this can be determined by knowledge of the process),  alterna-
tive methods, subject to the approval of the Administrator,
 U.  S.  Environmental Protection Agency, are required.

-------
                                1-152
     Filterable particulate matter may be determined along with S03 and
S02 (subject to the approval of the Administrator) by inserting a heated
glass fiber filter between the probe and isopropanol impinger (see
Section 2.1 of Method 6).  If this option is chosen, participate
analysis is gravimetric only; H^SCh acid mist is not determined separately.
2.  Apparatus
     2.1  Sampling.  A schematic of the sampling train used in this
method is shown in Figure 8-1; it is similar to the Method 5 train
except that the filter position is different and the filter holder does
not have to be heated.  Commercial models of this train are available.
For those who desire to build their own, however, complete construction
details are described in APTD-0581.  Changes from the APTD-0581 document
and allowable modifications to Figure 8-1 are discussed in the following
subsections.
     The operating and maintenance procedures for the sampling train are
described in APTD-0576.  Since correct usage is important in obtaining
valid results, all users should read the APTD-0576 document and adopt
the operating and maintenance procedures outlined in it, unless otherwise
specified herein.  Further details and guidelines on operation and
maintenance are given in Method 5 and should be read and followed
whenever they are applicable.
     2.1.1  Probe Nozzle.  Same as Method 5, Section 2.1.1.
     2.1.2  Probe Liner.  Borosilicate or quartz glass, with a heating
system to prevent visible condensation during sampling.  Do not use
metal probe liners.
     2.1.3  Pitot Tube.  Same as Method 5, Section 2.1.3.

-------
1-153

-------
                             1-154
     2.1.4  Differential Pressure Gauge.   Same as Method 5, Section
2.1.4.
     2.1.5  Filter Holder.   Borosilicate  glass, with a glass frit
filter support and a silicone rubber gasket.   Other gasket materials,
e.g., Teflon or Viton, may be used subject to the approval of the
Administrator.  The holder design shall  provide a positive seal  against
leakage from the outside or around the filter.  The filter holder shall
be placed between the first and second impingers.  Note:  Do not heat
     •
the filter holder.
     2.1.6  Impingers--Four, as shown in  Figure 8-1.  The first and
third shall be of the Greenburg-Smith design  with standard tips.  The
second and fourth shall be of the Greenburg-Smith design, modified by
replacing the insert with an approximately 13 millimeter (0.5 in.)
ID glass tube, having an unconstricted tip located 13 mm (0.5 in.)
from the bottom of the flask.  Similar collection systems, which have
been approved by the Administrator, may be used.
     2.1.7  Metering System.  Same as Method  5, Section 2.1.8.
     2.1.8  Barometer.  Same as Method 5, Section 2.1.9.
     2.1.9  Gas Density Determination Equipment.  Same as Method 5,
Section 2.1.10.
     2.1.10  Temperature Gauge.  Thermometer, or equivalent, to measure
the temperature of the gas leaving the impinger train to within 1°C
(2°F).
     2.2  Sample Recovery.
     2.2.1  Wash Bottles.  Polyethylene or glass, 500 ml. (two).
     2.2.2  Graduated Cylinders.  250ml, 1 liter.  (Volumetric flasks
may also be used.)

-------
                         1-155
     2.2.3  Storage Bottles.   Leak-free polyethylene bottles,
1000 ml  size (two for each  sampling run).
     2.2.4  Trip Balance.   500-gram capacity,  to measure to +0.5 g
(necessary only if a moisture content analysis 1s to be done).
     2.3  Analysis.
     2.3.1  Pipettes.  Volumetric 25ml, 100ml.
     2.3.2  Burette.  50 ml.
     2.3.3  Erlenmeyer Flask.  250 ml. (one for each sample blank
and standard).
     2.3.4  Graduated Cylinder.  100ml.
     2.3.5  Trip Balance.   500 g capacity, to  measure to +0.5  g.
     2.3.6  Dropping Bottle.   To add indicator solution, 125-ml  size.
3.  Reagents
     Unless otherwise indicated, all reagents  are to conform to  the
specifications established  by the Committee on Analytical Reagents of
the American Chemical Society, where such specifications are available.
Otherwise, use the best available grade.
     3.1  Sampling.
     3.1.1  Filters.  Same  as Method 5, Section 3.1.1.
     3.1.2  Silica Gel.  Same as Method 5, Section 3.1.2.
     3.1.3  Water.  Deionized, distilled to conform to ASTM specification
Dll93-74, Type 3.  At the option of the analyst, the KMnO^ test  for
oxidizable organic matter may be omitted when  high concentrations of
organic matter are not expected to be present.

-------
                                1-156
     3.1.4  Isopropanol, 80 Percent.  Mix 800 ml of isopropanol with 200
ml of deiom'zed, distilled water.
     Note:  Experience has shown that only A.C.S. grade isopropanol is
satisfactory.  Tests have shown that isopropanol obtained from commercial
sources occasionally has peroxide impurities that will cause erroneously
high suIfuric acid mist measurement.  Use the following test for
detecting peroxides in each lot of isopropanol: Shake 10 ml of the
isopropanol with 10 ml of freshly prepared 10 percent potassium iodide
solution.  Prepare a blank by  similarly treating 10 ml of distilled
water.  After 1 minute, read the absorbance on a spectrophotometer at
352 nanometers (Note:   Use a 1 cm path length).  If the absorbance
exceeds 0.1, the isopropanol shall  not be used.  Peroxides may be removed
from isopropanol by redistilling, or by passage through a column of
activated alumina.  However, reagent-grade isopropanol with suitably low
peroxide levels is readily available from commercial sources; therefore,
rejection of contaminated lots may be more efficient than following the
peroxide removal procedure.
     3.1.5  Hydrogen Peroxide, 3 Percent.  Dilute 100 ml  of 30 percent
hydrogen peroxide to 1 liter with deionized, distilled water.  Prepare
fresh daily.
     3.1.6  Crushed Ice.
     3.2  Sample Recovery.
     3.2.1  Water.  Same as 3.1.3.
     3.2.2  Isopropanol, 80 Percent.  Same as 3.1.4.
     3.3  Analysis.
     3.3.1  Water.  Same as 3.1.3.

-------
                                1-157
     3.3.2  Isopropanol, 100 Percent.
     3.3.3  Thorin Indicator.   l-(o-arsonophenylazo)-2-naphthol-3,
6-disulfonic acid, disodium salt, or equivalent.   Dissolve 0.20 g
in 100 ml of deionized, distilled water.
     3.3.4  Barium Perchlorate (0.0100 Normal).   Dissolve 1.95 g of
barium perchlorate trihydrate (BalClO.L-SHpO)  in 200 ml  deionized,
distilled water, and dilute to 1 liter with isopropanol;  1.22 g of
barium chloride dihydrate (BaCU^HpO) may be used instead of the
barium perchlorate.  Standardize with sulfuric  acid as in Section 5.2.
This solution must be protected against evaporation at all times.
     3.3.5  Sulfuric Acid Standard (0.0100 N).   Purchase  or standardize
to +0.0002 N against 0.0100 N NaOH that has previously been standardized
against primary standard potassium acid phthalate.
4.  Procedure
     4.1  Sampling.
     4.1.1  Pretest Preparation.  Follow the procedure outlined in
Method 5, Section 4.1.1; filters should be inspected, but need not  be
desiccated, weighed, or identified.  If the effluent gas  can be
considered dry, i.e., moisture free, the silica gel need  not be weighed.
     4.1.2  Preliminary Determinations.  Follow the procedure outlined
in Method 5, Section 4.1.2.
     4.1.3  Preparation of Collection Train.  Follow the  procedure
outlined in Method 5, Section 4.1.3 (except for the second paragraph
and other obviously inapplicable parts) and use Figure 8-1 instead  of
Figure 5-1.  Replace the second paragraph with:   Place 100 ml of

-------
                                 1-158
80 percent isopropanol in the first impinger,  100 ml  of 3 percent
hydrogen peroxide in both the second and third impingers; retain a
portion of each reagent for use as a blank solution.   Place about
200 g of silica gel in the fourth impinger.
     Note:  If moisture content is to be determined by impinger analysis,
weigh each of the first three impingers (plus  absorbing solution) to
the nearest 0.5 g and record these weights. The weight of the silica
gel (or silica gel plus container) must also be determined to the nearest
0.5 g and recorded.
     4.1.4  Pretest Leak-Check Procedure.  Follow the basic procedure
outlined in Method 5, Section 4.1.4.1, noting  that the probe heater
shall be adjusted to the minimum temperature required to prevent
condensation, and also that verbage such as, ". . . plugging the
inlet to the filter holder . . .  ," shall be replaced by, ". . . plugging
the inlet to the first impinger . . . ."  The  pretest leak-check is
optional.
     4.1.5  Train Operation.  Follow the basic procedures outlined in
Method 5, Section 4.1.5, in conjunction with the following special
instructions.  Data shall be recorded on a sheet similar to the one
in Figure 8-2.  The sampling rate shall not exceed 0.030 m /min
(1.0 cfm) during the run.  Periodically during the test, observe the
connecting line between the probe and first impinger for signs of
condensation.  If it does occur, adjust the probe heater setting
upward to the minimum temperature required to  prevent condensation.
If component changes become necessary during a run, a leak-check shall

-------
                                                                   1-159
                           U


                           CC
 c


 o
      CC  UJ  CC
<  O.  —I  0.
                                  CC
                                          o
                                          CJ
   z
                 <  =
                 O  E
                                      CC
                                      o
                     UJ   UJ  U.

                     Z   S  o
UJ OC EC
CC O UJ
UJ u. <
JJ S 0
2 J z c/>
"j o<
^ O -J
UJ
oc
t- oc
^ UJ
UJ
OL OC
E O
i/>
0
UJ U.
_JO_
0°

*~?

UJ
III
c
/> ° E
ts
—1
uj uj o EC "J E e
ec u. < o 5 Eo
& u.
5
>
O uj <3

o —
f!
UJ
CC
3
CO £L "
UJ
^
3
o
z
Q.
I/)





S .E
c —
c
s e
—
z
2s
Uj 00
^ Z
oc
*"








































































































































































































































































































































en


01
a
















i
O







I














UJ
oc

"*
                                                                                                                                          CD

                                                                                                                                          0)
                                                                                                                                          5
                                                                                                                                          o>

                                                                                                                                          u.

-------
                                1-160
be done immediately before each change, according to the procedure
outlined in Section 4.1.4.2 of Method 5 (with appropriate modifica-
tions, as mentioned in Section 4.1.4 of this method); record all
leak rates.  If the leakage rate(s) exceed the specified rate, the
tester shall either void the run or shall  plan to correct the sample
volume as outlined in Section 6.3 of Method 5.  Immediately after
component changes, leak-checks are optional.  If these leak-checks
are done, the procedure outlined in Sp-.tion 4.1.4.1  of Method 5
(with appropriate modifications) shall be  used.
     After turning off the pump and recording the final readings  at
the conclusion of each run, remove the probe from the stack.  Conduct
a post-test (mandatory) leak-check as in Section 4.1.4.3 of Method 5
(with appropriate modifications) and record the leak rate.  If the
post-test leakage rate exceeds the specified acceptable rate, the
tester shall either correct the sample volume, as outlined in
Section 6.3 of Method 5, or shall void the run.
     Drain the ice bath and, with the probe disconnected, purge the
remaining part of the train, by drawing clean ambient air through the
system for 15 minutes at the average flow rate used for sampling.
     Note:  Clean ambient air can be provided by passing air through
a charcoal filter.  At the option of the tester, ambient air (without
cleaning) may be used.
     4.1.6  Calculation of Percent Isokinetic.  Follow the procedure
outlined in Method 5, Section 4.1.6.

-------
                               1-161
     4.2  Sample Recovery.
     4.2.1  Container No.  1.   If a moisture content analysis is to
be done, weigh the first impinger plus contents to the nearest 0.5 g
and record this weight.
     Transfer the contents of the first impinger to a 250-ml graduated
cylinder.  Rinse the probe, first impinger, all connecting glassware
before the filter, and the front half of the filter holder with 80 per-
cent isopropanol.  Add the rinse solution to the cylinder.  Dilute to
250 ml with 80 percent isopropanol.  Add the filter to the solution,
mix, and transfer to the storage container.  Protect the solution against
evaporation.  Mark the level  of liquid on the container and identify the
sample container.
     4.2.2  Container No.  2.   If a moisture content analysis is to be
done, weigh the second and third impingers (plus contents) to the nearest
0.5 g and record these weights.  Also, weigh the spent silica gel (or
silica gel plus impinger)  to the nearest 0.5 g.
     Transfer the solutions from the second and third impingers to a
1000-ml  graduated cylinder.  Rinse all connecting glassware (including
back half of filter holder) between the filter and silica gel impinger
with deionized, distilled water, and add this rinse water to the cylinder.
Dilute to a volume of  1000 ml with deionized, distilled water.  Transfer
the solution to a storage container.  Mark the level of liquid on the
container.  Seal and identify the sample container.
     4.3 Analysis.
     Note the level of liquid in containers 1 and 2, and confirm whether
or not any sample was  lost during shipment; note this on the analytical
data sheet.  If a noticeable amount of leakage has occurred, either void
the sample or use methods, subject to the approval of the Administrator,
to correct the final results.

-------
                                1-162
     4.3.1  Container No.  1.   Shake the container holding the
isopropanol solution and the  filter.   If the filter breaks up,
allow the fragments to settle for a few minutes before removing
a sample.  Pipette a 100-ml  aliquot of this solution into a 250-ml
Erlenmeyer flask, add 2 to 4  drops of  thorin indicator, and titrate
to a pink endpoint using 0.0100 N barium perchlorate.   Repeat the
titration with a second aliquot of sample and average  the titration
values.  Replicate titrations must agree within 1 percent or 0.2 ml,
whichever is greater.
     4.3.2  Container No.  2.   Thoroughly mix the solution in the
container holding the contents of the  second and third impingers.
Pipette a 10-ml aliquot of sample into a 250-ml Erlenmeyer flask.
Add 40 ml of isopropanol,  2 to 4 drops of thorin indicator, and
titrate to a pink endpoint using 0.0100 N barium perchlorate.  Repeat
the titration with a second aliquot of sample and average the titration
values.  Replicate titrations must agree within 1 percent or 0.2 ml,
whichever is greater.
     4.3.3  Blanks.  Prepare  blanks by adding 2 to 4 drops of thorin
indicator to 100 ml of 80 percent isopropanol.   Titrate the blanks  in
the same manner as the samples.
5.  Calibration
     5.1  Calibrate equipment using the procedures specified in the
following sections of Method  5:  Section 5.3 (metering system); Section 5.5
(temperature gauges); Section 5.7 (barometer).   Note that the recommended
leak-check of the metering system, described in Section 5.6 of Method 5,
also applies to this method.
     5.2  Standardize the barium perchlorate solution  with 25 ml of
standard sulfuric acid, to which 100 ml of 100 percent isopropanol  has
been added.

-------
                              1-163
6.  Calculations
     Note:  Carry out calculations retaining at least one extra
decimal figure beyond that of the acquired data.   Round off figures
after final  calculation.
     6.1  Nomenclature.
                                               2     2
     A      = Cross-sectional area of nozzle, m  (ft ).
     B      = Water vapor in the gas stream, proportion by volume.
     C,, cn  = Sulfuric acid (including S00) concentration, g/dscm
      H2bU4                               6
              (Ib/dscf).
     CSQ    = Sulfur dioxide concentration, g/dscm (Ib/dscf).
     I      = Percent of isokinetic sampling.
     N      = Normality of barium perchlorate titrant, g equivalents/
              liter.
     P.      = Barometric pressure at the sampling site, mm Hg  (in.  Hg)
     P      = Absolute stack gas pressure, mm Hg  (in. Hg).
     P  d   = Standard absolute pressure, 760 mm  Hg (29.92 in.  Hg).
     T      = Average absolute dry gas meter temperature (see
              Figure 8-2), °K (°R).
     T      = Average absolute stack gas temperature (see Figure  8-2),
              °K (°R).
     Tstd   = standard absolute temperature, 293°K (528°R).
     Vfl     = Volume of sample aliquot titrated,  100 ml for  HpSO,
              and 10 ml  for S0?.
     Vlc    = Total volume of liquid collected in impingers  and
              silica gel, ml.

-------
                                1-164
     V      s Volume of gas sample as measured by .dry gas meter,
              dcm  (dcf).
     V  / td)= Volume of gas sample measured by the dry gas meter
              corrected to standard conditions, dscm (dscf).
     v      = Average stack gas velocity, calculated by Method 2,
              Equation 2-9, using data obtained from Method 8,
              m/sec (ft/sec).
     V  •,   = Total volume of solution in which the sulfuric acid
      soln
              or sulfur dioxide sample is contained, 250 ml or
              1000 ml, respectively.
     V.     = Volume of barium perchlorate titrant used for the
              sample, ml.
     V^.    = Volume of barium perchlorate titrant used for the
      tb
              blank, ml.
     Y      - Dry gas meter calibration factor.
     AH     = Average pressure drop across orifice meter, mm (in.)
              H20.
     8      - Total sampling time, min.
     13.6   = Specific gravity of mercury.
     60     = sec/min.
     100    = Conversion to percent.
     6.2  Average dry gas  meter temperature and  average  orifice
pressure drop.   See data sheet (Figure 8-2).
     6.3  Dry Gas Volume.   Correct the sample  volume  measured  by
the dry gas meter to standard  conditions  (20°C and 760 mm Hg or
68°F and 29.92 in.  Hg)  by  using Equation  8-1.

-------
                                1-165
                      Pbar * (T>   . K  „ Y      * "H/13'6)
                      —    -- KlV
                                                      Equation 8-1
where:
     KI = 0.3858 °K/mm Hg for metric units".
        = 17.64 °R/in. Hg for English units.
     Note:  If the leak rate observed during any mandatory leak-checks
exceeds the specified acceptable rate, the tester shall either correct
the value of V  in Equation 8-1 (as described in Section 6.3 of
Method 5), or shall invalidate the test run.
     6.4  Volume of Water Vapor and Moisture Content.  Calculate the
volume of water vapor using Equation 5-2 of Method 5; the weight of
water collected in the impingers and silica gel  can be directly
converted to milliliters (the specific gravity of water is 1  g/ml).
Calculate the moisture content of the stack gas, using Equation 5-3
of Method 5.  The "Note" in Section 6.5 of Method 5 also applies to
this method.  Note that if the effluent gas stream can be considered
dry, the volume of water vapor and moisture content need not  be calcu-
lated.
     6.5  Sulfuric acid mist (including SO-) concentration.
                    N (V.  - V.,) ,  -nn-
     Cu ,n   =  K9      l    tD  \  va  /            Equation 8-2
       74      *-          \i
       2  4                 Vm(std)
where:
     K2 = 0.04904 g/mi Hi equivalent for metric units.
        = 1.081 x 10"4 Ib/meq for English units.

-------
                               1-166


     6.6  Sulfur dioxide concentration

                   N f\i  _ v  )  I _
                   H ^Vt   vtb;  I   V
     Ccn   =  K,   	n	——               Equation 8-3
      bU2      J          m(std)
where:
     K-  = 0.03203 g/meq for metric units.
                     -5
         = 7.061 x 10   Ib/meq for English units.

     6.7  Isokinetic Variation.

     6.7.1  Calculation from raw data.
                           (VmY/Tm) (Pfc.. + AH/13.6)]
                       60 e v  P  An


                                                      Equation 8-4
where:
     K4 = 0.003464 mm Hg-m3/ml-°K for metric units.

        = 0.002676 in. Hg-ft3/ml-°R for English units.

     6.7.2  Calculation from intermediate values.
              Ts Vstd) pstd
           Tstd vs * An    60
          K
           5  Ps y
where:
     Kc = 4.320 for metric units.
      o

        = 0.09450 for English units.

-------
                               1-167
     6.8  Acceptable Results.  If 90 percent <_ I <_ 110 percent,
the results are acceptable.  If the results are low in comparison
to the standards and I is beyond the acceptable range, the
Administrator may opt to accept the results.  Use Citation 4 in
the Bibliography of Method 5 to make judgments.  Otherwise, reject
the results and repeat the test.
7.  Bibliography
     1.  Atmospheric Emissions from Sulfuric Acid Manufacturing
Processes.  U. S. DHEW, PHS, Division of Air Pollution.  Public
Health Service Publication No. 999-AP-13.  Cincinnati, Ohio.  1965.
     2.  Corbett, P. F.  The Determination of SO,, and S03 in Flue
Gases.   Journal of the Institute of Fuel.  24_:237-243.  1961.
     3.  Martin, Robert M.  Construction Details of Isokinetic
Source Sampling Equipment.  Environmental Protection Agency.  Research
Triangle Park, N. C.  Air Pollution Control Office Publication No.
APTD-0581.  April, 1971.
     4.  Patton, W. F.  and J. A. Brink, Jr.  New Equipment and
Techniques for Sampling Chemical Process Gases.  Journal  of Air
Pollution Control Association.  1_3:162.  1963.
     5.  Rom, J. J.  Maintenance, Calibration,  and Operation of
Isokinetic Source-Sampling Equipment.  Office of Air Programs,
Environmental Protection  Agency.  Research Triangle Park, N. C.
APTD-0576.  March, 1972.
     6.  Hamil, H. F.  and D. E. Camann.  Collaborative Study of
Method for Determination  of Sulfur Dioxide Emissions from Stationary

-------
                              1-168

Sources (Fossil Fuel-Fired Steam Generators).   Environmental  Protection
Agency.  Research Triangle Park, N. C.   EPA-650/4-74-024.   December,
1973.
     7.  Annual Book of ASTM Standards.   Part  31; Water, Atmospheric
Analysis,   pp. 40-42.  American Society for Testing and Materials.
Philadelphia, Pa.  1974.

-------
FIELD DATASHEETS





     PART II

-------
                     TRAVERSE POINT LOCATION FOR CIRCULAR DUCTS
PLANT	
DATE	
SAMPLING LOCATION 	
INSIDE OF FAR WALL TO
  OUTSIDE OF NIPPLE, (DISTANCE Ai _
INSIDE OF NEAR WALL TO
  OUTSIDE OF NIPPLE, (DISTANCE B) _
STACK I.D., (DISTANCE A - DISTANCE B).
NEAREST UPSTREAM DISTURBANCE	
NEAREST DOWNSTREAM DISTURBANCE _
CALCULATOR	
SCHEMATIC OF SAMPLING LOCATION
TRAVERSE
POINT
NUMBER
























FRACTION
OF STACK I.D.
























STACK I.D.
























PRODUCT OF
COLUMNS 2 AND 3
(TO NEAREST 1 '8 INCH)
























DISTANCE B
























TRAVERSE POINT LOCATION
FROM OUTSIDE OF NIPPLE
(SUM OF COLUMNS 4 & 5)
























                                            II-l

-------
                METHOD 2 GAS  VELOCITY  AND VOLUME DATA
PLANT AND CITY

RUN DATE
I
/
I
/
I
SAMPLING LOCATION

CLOCK
TIME
RUN
NUMBER

OPERATOR

AMB . TEMP .
(°F)

BAR. PRESS
(in. Hg)

STATIC PRESS
(in. H20)

MOLECULAR
WT.
i i i i
STACK INSIDE DIMENSION (in.)
DIAM OR SIDE 1
i i i i i
SIDE 2
i i i i i _.
PITOT
TUBE Cp
t i i
          FIELD DATA
TRAVERSE
POINT
NUMBER





























POSITION
(in.)





























VELOCITY
HEAD
(Ap ),
in. H20





























STACK
TEMP. , CF





























CYCLONIC FLOW DETERMINATION
Ap at
0° REFERENCE





























AVERAGE ANGLE (=)
ANGLE (oc)
WHICH YIELDS
A NULL Ap






























Average of  ( a) must be less than  10 degrees  to  be  acceptable.



                                II-2

-------
      METHOD 2  PITOT TUBE CALIBRATION
PITOT TUBE IDENTIFICATION NUMBER:.

CALIBRATED BY:	
DATE:.

RUN NO.
1
2
3
"A" SIDE CALIBRATION
APstd
em H20
(in. H20)




*P(s)
cm H20
(in. H20)



AVERAGE
CP(SJ





DEV.





RUN NO.
1
2
3
"B" SIDE CALIBRATION"
Apstd
em HpO
|m.H20)




*P(s)
cm H20
(in. H20)



AVERAGE
CP(S)





OEV.




  DEV.-  .Cp(S)-Cp(s)(avq.)  (MUST BE £0.01)
  Cp DIFFERENCE: Aavg-Bavj*	(MUST BE £ 0.01)
  NOTE:  Not required if  pitot tube
          meets geometry standards.
                        II-3

-------
Date
              Integrated Bag Sampling Field Data




                      Run No.  	






                        Plant
Sampling location
Barometric pressure



Ambient temp. °F 	



Operator 	
                      11  Hg.
             Stack  temp.  °F
Sample bag leak checked
          Sample train leak checked
Time




























Traverse
point




























Rate weter
flow rate,
Q cm /min.




























Average
n _ n
% Dev.a





























Fyrite,
°2




























co2





























  % Dev. =  (-
-) 100
                              II-4
(Must be  -10%)

-------
o
(__
<:
oc
UJ
I—
LU
O
CD

UJ
o
UJ
_i
o
o:
a
o
CJ

















o
z







LU
H-

























§
O
_J
o







G
33
Z
h-
Z
O
o
o
LU
1—
CL
U.
t—
^
f Z
,1 O CJ
CM r; <
















































LU
cc
Q =3
• g ^
i— tr
"~^ — r ^ LU LU
LU 75 ~ 3= Q.
*• 0 £ * =
r; ~j Q- -= LU
^"~ — * ^^
C3 C3 H"






















Q
LU
UJ
IE
^— '
UJ
P; i— a: _i
— •_ o
^ ^ LU >- z fc •-
_j — j — i *•; LU s <:
£i£i«£££
< 4* ^ «f Z £ Q> , — .
0^tloDl«S
*~ 3
^_ O
LU >: E
* § a
«£ i^> -0
_l 
„











i—
LU
Z
<_) LU
t—
LU
Z
33 Q
f— <
O LU

0 =3
^ — ^g"
 O
^3 5
i ^

VI tj
CM O








|—
O





















                                             II-5

-------
J
U
u o o>
,J U
V)
IMPINGER
VOLUME ,
ml







3
U.





H
Z
M
So  -
M Z O
OC    Z
U X
EH Q X S
< ffi O Z
Z    0 -
  u
u J a: os
flQ p^ UJ UJ
O E EH EH
oc *cl u u
a, ui SE 2
— < u
>• 2 z
     kH
   X Ct
J < 3
< ui a
t. J

K^i
Ul < D

S5^
X U, >
  z a
  o o
fH EH <   Z U
z < c2 u    M
  EH Ul 1«
   ' Ul —
     y H

     a. s
       EH
  Ul U U
  EH M Z
     QS ui
     H J
O Z W
Z U £ Ul
     " 03

0!
SOS U
o u
i-1 m o « 2 £
< «S z M a, «
OS U M V) Z ^-
Sb.5.SM.u
2 O W Q H
W >J
Z VI
* ,85
I
i
a i
e '
4 «
PC U
U EH
W
u o
0. ><
x a.
< Q
Ul
M <




Ou
fn O

li 3
O O
E
_
•
^^
H O
3 *
z "c
M -H
• B-u
m
E
td

o
J 	
X UX^t
« Z I Z w
U M < D *4
EH Q W J —
W *^I O^""*
o
H) «
<
J
W W EH O
2 EH W IN
a z w S • s
Ul U Ul . 	
ui 2 o u i: .
w u a <
a: u. o •
j o c
-4 ^- »H
a. u< < u. —
Q 5 §
o


1-

tc fn h,
i
( a. —
•H Ul U
ui a, o

u
U
Z -Z
< 
                                                                                                                         O
                                                                                                                         u
                                                                                                                         (D
                                                                                                                         0)
                                                                                                                         e
                                                                                                                         m
                                                                                                                         c
                                                                                                                         (0
                                                                                                                         
-------

















.
0
2

C
O
•H
IB
O
•H
MH MH
O -H
4J





C
01
TJ
M

0)
4-1 rH
01 N
OI N
-C 0
t/) 2
C
•H

1

















14
U
4J
0>
g
MH
O

C
g
\
ro
£





01
4J
IB
«
g
(8 • X
•H 0 ifl
Q 2 Oi
rJ
0)
l-H g r-H
N rH IB
N 01 C
o .c -H
2 EH fc




Ol

C
•H
CP
.* s
o
01 g
J3 g
CJ

i^
IB
01
rJ

CP
c
•rH
lj
3
Q

g
3
3
0
IB
>









.
O
2

14
01
4-1
rH
•H
ft,





>r






c







o

4-1 "~h.
IB CJ





r— 1
IB





CP
c





0)
1 .
CP
c
•rH
&
X
s.
04
0)
.44-133
01 4-> 4-1 01
-U 0) IB 01
 O O O XI M
01 -rt V-l H U g (B
2 04 0* Q* O4 < CQ





3

OI
M
•3
4J
01
•rH
O
s

T3
0)

3
01
in
<
4-1
c
us
    4-)

>i  18
4-1  O

H  O

CJ  rj
        C  ,4
        O  O
       •H  4->
           o  a   a;
o
IB
4J
in




































O
2 ••



















01
14 X
01 H
4-1 (B
r-H g
•H 01
r fV
MH K


^•v . 	






















O
(N
X
c
•H
^~
o
IN
X

g
g
















^'
01
04
' 	 '
01
M 01
3 rH
•U 3
01 0)
•rH 01
O 01
S rH
o
CM
X
q
•H
^
O
(N
X

g
g





























04
<:
m
a, >H o
T3 O C
01 O 4J 0)
g -H O .H
3 4-1 IB 0)
01 IB &-i UH
Ul 4-1 01
< U2 CJ 05











































,















0
2 0
2


















O X H M
a 0) o)
g 4J 4J
IB 0) 0)
W 3E S

M
01
4->
rH
•rH
[14
rH
0) 0
)H
3 01 CT> rH
4-1 iB C 01
 C
Ol MH IB 01
a o 0) -a
g rH R
oi o
EH U
1
a >i
g .H
01 13
4J M
4-1 01
01 IB 4-1
rH 01
CX 0) g
g rl
iB 3 01
01 4J IB
IB CP
01 rH
IB 0)
CJ)
01
rH
' h
• o
a.—
g
0) CJ
4-1 0
M
01
en

•H 5T
0,0

•H
U
4-1 a
01
IB
rH
4J ST
01 0
rH — '
4J
3 CJ
O o

^^^
• h
4-> 0
o> —
rH
C CJ
H 0
1=
^ f-*
Q< — n
gi i
IB
01

01
IB
O
01 1 X
lH >H rH 01 01 <]
3 0) n) in u —
01 *4 -rH O -rl
0] MH 4-1 VH ^ V-i
oi -rt q o -H 0)
r-l T3 01 (B .H 4J
a, o 01
E
>1
4-1
•H T3 -~
O IB V
o 01 a.
rH £ 


1
IB
.* -H 01
U 01 r-l
IB a 3
4-1 g 4J
c/i oi
4J
s
3
3
u
IB
>
.X
o
0
r-H
u
CP
c
•H
rH
a
g
IB
W


0)
01
rH
01
>
iB
W
EH
01 MH
g —
3
rHn
O g
>
0
O H
0) J3
g
•H T
4-1 (N
c
-•H
01 g
•S .
•P ^~-
cr>
— •



r4
•P 01
q xi
•H g
O 3
a q














































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































































*
X
-?


.
01
>
<




,
tp
>
<




r-H
IB
4J
0
EH























.
tT
>
<



.
X
fl








r-H

-------
                         METHOD  5

       PARTICULATE SAMPLE RECOVERY AND INTEGRITY SHEET

Plant: 	 Sample date: 	
Sample location: 	 Run no.: 	
Sample recovery person: 	 Recovery date: 	
Filter (s) no.: 	
                          MOISTURE
Impingers                          Silica gel
Final volume  (wt) 	 ml (gm) Final wt.	
Initial volume  (wt) 	 ml(gm) Initial wt. 	
Net volume  (wt) 	 ml (gm) Net wt.	
     Total moisture 	„
Color of silica gel 	
Description of impinger water 	
                      RECOVERED SAMPLE
Filter container no.                 sealed
Description of particulate on filter
Acetone rinse                   Liquid level
container no.                   marked
Acetone blank                   Liquid level
container no. 	 marked
Samples stored and locked 	
Remarks:
Date of laboratory custody 	
Laboratory personnel taking custody
Remarks:
                            II-8

-------
                   BLANK ANALYTICAL DATA
Plant
Sample location
Relative humidity

Type of blank 	
Liquid level at mark and container sealed

Density of blank (pa) 	

Blank volume (Va) 	

Date and time of wt. 	

Date and time of wt.
                                         ml

                                         Gross wt.

                                         Gross wt.
                                 Average gross wt.

                                          Tare wt.

                              Weight of blank (ma)
          Ca =
                 ma
               Va pa
T(
                              mg

                              mg

                              mg

                              mg

                              mg
Note:  In no case shall a blank residue greater than  (0.01
mg/g) or 0.001% of the weight of blank used be subtracted
from the sample weight.
Remarks:
          Signature of analyst

          Signature of reviewer
                             II-9

-------
         METHOD 5 TRAIN ANALYTICAL PARTICULATE DATA
Plant 	

Sample location
Relative humidity
Density of acetone (pa)
Run No.
       g/ml
Sample
type
Acetone rinse
filter (s)
Sample
identifiable


Liquid level marked
and/or container sealed


Acetone rinse container no.

Acetone rinse volume (Vaw)
    ml
Acetone blank residue concentration (Ca)
Wa = Ca Vaw pa = (
Date and time of wt
Date and
Filter (s)
Date and
Date and
time of wt
Weight of
container
time of wt
time of wt
Weight
Weight of
) ( ) ( ) =
Gross wt
Gross wt
Average gross wt
Tare wt
Less acetone blank wt (Wa)
particulate in acetone rinse
no .
Gross wt
Gross wt
Average gross wt
Tare wt
of particulate on filter (s)
particulate in acetone rinse
Total weight of particulate
mg/g
mg
mg
mg
mg
mg
mg
mg
mg
mg
mg
mg
mg
mg
mg
Note: In no case shall a blank residue greater than  (.01 mg/g) or
.001% of the weight of acetone used be subtracted from the sample
weight.

Remarks:
     Signature of analyst

     Signature of reviewer
                                11-10

-------
Date
                    NOZZLE CALIBRATION
                     Calibrated by

Nozzle
identification
number

D-j^, in.

Dp, in.

D.,, in.

AD, in.

D
avg

where :
D-, „ .. = nozzle diameter measured on a different diameter, in.
' ' ' Tolerance = measure within 0.001 in.
AD = maximum difference in any two measurements, in.
          Tolerance = 0.004 in.
   D
    avg
= average of D,, D_, and D_,
                         Nozzle calibration data.
                          11-11

-------
    CALIBRATION

    DRY GAS METER & ORIFICE

    DATE	

    BAROMETRIC PRESSURE, Pfa
                          METER BOX NO.
.in. Hg.
DRY GAS METER NO.
Orifice
manometer
setting
AH
in. H20
0.5
1.0
1.5
2.0
3.0
4.0
Gas volume
wettest
meter
V
ft3
5
5
10
10
10
10
Gas volume
dry gas
meter
V
ft3












Wet test
Meter
V
°F












Dry gas meter
Inlet
V
°F












Outlet
V
°F












Average
V
°F






Time
e,
min.






Y






AH@






                                                                                   Average

AH


0.5
1.0
1.5
2.0
3.0
4.0

AH
13.6

0.0368
0.0737
0.110
0.147
0.221
0.294
Y
Vw Pb 

-------
H
^
<
u

CJ
w
W
O
(Xi
               £
               Q.
               o>
               4-1
               o>
               i.
               ex
     O)
     L.
     3
     4/1
     t/1

     £
     Q.
          1-    O
          4-1   CO
                    1-
                    «J
Ol
4->
>0
                .
               O>
               *•    >>
               1)    U
               x   o


>-











"o^
t£>
s '*
>O 4-
*?
X
4- 4J^
^ •* •*
^ si".
^ <•£
a.
X
ex
•o
2»
>-











CTi
EC cn
3 ••- I
3 4->
U 4-» c
10 OJ —
> U1
c:
.§ ^
t— *
O










1

CO
m
CT

>•
O





W
O
4->
4_)
Ol
3E

«
ID
cn

> X
L.
(/
(C
(J
1
c
>
t/1
*«
^

a
c.

U-

t
T3
tJ
aj
4-1

4-J
HI
i
t.
OJ
4-*
Of
E
o
c
og




o
QJ
CT!
fl - U-
i- -a o
QJ 4->
>
<
4->
a> *
^- O Lu
4-> T) 0
3 4-»
0


4->
a> -
r— •-— U_
C T3 o
>— i +->
t_
OJ " U.
4-> i 0
(U -4-1



t.

OJ > «*-

l_
QJ * m
4-> S +•»
m > i«_

*
o^ o
C * C\J
— X X
4-> <
+-> c
at •*-
t/»
•*-» •
u^ 0
QJ Z
H-














































0















































































O



























































o


















It
X
4->
Wl
QJ
*J
1
4-*
1/1
O
Q_












































4









-
«
i
•*
«

«
2
4
|
J
&
!
1
J
,»









M>
»*.
*
I
w
«
*-
b>
9
•rf
&
W
r
«*
£
i
1
j
v
»









t^.
•
t.
*rf

^
«

|
e
wt
3.
2
•
k
J
•








*
«
i
2
X
V
JJ!

O
i
«
*e
2
3
S
*—
V"







t*;
.
i
•*
>«
L.

2
^_
O
E
I
2
t
:
V
W
O
V
»
<*
i.
^
2
X
A
1
*
A
O
W
V
I
•*
>t
w.
T>
«t
1
c
w»
2
O
£
*
1
»
U
<
> V










°^
c
^
B
•1
W
£
b
o
<•>
w*
e
*
^
**
«
£
«• i
•O '
• »
2 :
? ;
&. •
5 .


















i
i
i
i
-— >
•
|












*»
«
^ 9
s r ;
O • •
* ' — 5
»- h.
s ?!
I I I
• a —
W 4
«J — tJ
Z - «-
! i •
!!!
A O
                                                     11-13

-------
Date:
                              Thermocouple No.:
Ambient Temperature:.

Calibrator:	
                           °C Barometric pressure:
."  Hg
                           Reference:  Mercury-in-glass:
                                                        Other:_
Reference
point
No.a
















Source
(specify)
















Reference
thermometer
temperature,
°C
















Thermocouple
potentiometer
temperature,
°C
















Q
Difference ,
%
















,  Every 30°C (50°F) for each reference point
  Type of Calibration system used
  (Ref. temp; "C + 273) - (Test therm, temp.
                  Ref. temp. CC + 273
                                                    >C + 273)
                                                             x 100<1.5%
                   Stack Temperature Sensor Calibration Data Sheet
                                11-14

-------


















































































































C^
^












05
0
U 2














0 < O
2 fc, M _
















_ j_| p_, ^
•U EH 2 < .5
4-1 EH O U g
^ W M O ^
6 tO EH J
<. Q
X OS « EH W
EH W CQ 2 O
• ^ tH r- 1 1— I U-l
O 2 < J O ID























2 Ctj- U < (X CM
J 3C U CO
(y U Cd &S
J CJ U OS hH J 3
SH W Dj CQ CQ W QJ CM rt|
EH EH 2 O O EH 2 2 2
rH
K
g
5
^



































1
05




































W j


































U



































X


































u
W J»H



































X U
U W
OS < S
o< n o a. x U
05 2 2 <
U W W W -*5
H EH X EH .J <
2 i-3 05 05 < O W
2 O EH 2 OQ EH rn1 r}
W EH W Z <
EH ) "^ irf 2£ CO Ctf fT1 ^ ^
^CUffiOCQWM£nrtJ
^swDnOE^cnHHz
h42cw*£0*wszw
C*COOca0.2rt|H&,

E ao
M E 0
0)
4J


.
a
E
UJ OJ
(04-1 —
en Cn
VH o
>i OJ^
JH 4J
Q Q) U
Eo


C
0
4J -H
C 4-> —
(U (8 +1
U -H —
^ >
flj 3) SP
a, a

OJ
E
3 —
O 0) 4->

0) —
(U 4-1
rH 
t3
CO <

Cn
3 C —
O -H E
rH 4J 4H
IM 4J U
0) —
0) tn
rH
a a) g
E -u a
tO M


Q) 0) —
*~H E?*^
a 3 -P
E rH UH
tj O ' — '
00 > rH

"£
^ J3
0 (I)
O E **
rH -rH (N
U EH —


Cn
C
•H
i-H ^
a a) c
£ E -H
(fl -H E
tO EH —































































































































































































































































































































































































































































































































































































•
a
£
0)
EH

•
X
(0
2







,
Cn
>



•
£>
0)
Q

.
X
fl
2





.
Cn
>
(0

s
^
 en a
<3 > EH
(T3 M
I ^
E
e >
"*< <

n
. —
of
—

•n
0
(0

^
0)
03

ij
C
0)
u

0)


_^^
+ 1
11-15

-------
PLANT
                  METHOD 6  SAMPLE RECOVERY  AND INTEGRITY DATA
SAMPLE LOCATION
                                 FIELD DATA CHECKS
SAMPLE RECOVERY PERSONNEL
PERSON WITH DIRECT  RESPONSIBILITY FOR RECOVERED SAMPLES
Sample
no.
1
2
3
4
5
6
Blank
Sample
identification
no.







Date of
recovery







Liquid
level
marked







Stored
in locked
container







REMARKS
SIGNATURE OF FIELD  SAMPLE  TRUSTEE
                              LABORATORY DATA CHECKS




LAB PERSON WITH DIRECT  RESPONSIBILITY FOR RECOVERED SAMPLES




DATE RECOVERED SAMPLES  RECEIVED	




ANALYST
Sample
no.
1
2
3
4
5
6
Blank
Sample
identification
no.







Date of
analysis







Liquid
at marked
level







Sample
identified







REMARKS
SIGNATURE OF LAB  SAMPLE TRUSTEE
                                       11-16

-------
                          METHOD 6

               SULFUR DIOXIDE ANALYTICAL DATA
PLANT
DATE
E LOCATION ANALYST
LITY OF
1
BARIUM PERCHLORATE 2
3
ml N
ml N N
ml N
Run
No.
1
2
3
4
5
6
Sample
No.






Total
Volume
of Sample






Sample
Aliquot






Volume of Titrant V , ml
1st
Titration






2nd
Titration






Average






Blank Analysis - Volume of titrant  1st titration.
                                    2nd titration.
                                    Average	
1st titration
2nd titration
= 0.99 to 1.01 or 1st titration - 2nd titration = 0.2 i
Signature of Analyst	
Signature of Reviewer or Supervisor.
                            11-17

-------















































0<
•
c
•H


u




A3
•Q
CL
^^

V
3
n
M
0)
Ij
a.

u
u
V
E
>t V O
*J 4J 14
O Q ffl




































c














0
• H
iJ
IQ
o
o




















-H kl
O
Q) &J
4J ^ A3
c a u
i E ai
•-H *a o.
a in O





hi
3
Li
01
a
01
^4
4J
C
t-t

4)
U
3
0)
M
4)
ft. X

^4 •
V C
•H -^
JJ
•H
e




JQ
•H
Of
•


•H
4J

U.
0
•a
ft.




^
a
cr

ij


•M
<
cr
•^



«
§ "3 ^
"o^ > "e
o c
01
*fl ^ 0
•-> z
h. T3
C
Ifl
3
O I- b.
*r
E
4J
01
-H 01 t,
a E £
E H I
nj *J v
tf)  o

0)
li
10































































































































































































































•M«
•H
A

•f

••4 a.
^ •
•» 0
VO
1 »

5 •-«
a. *J
i i
a. H
« A
                                          a)
                                          
-------
Plant
                                       Date
Sample Recovery Personnel
      Barometric  Pressure, (P   )
                            ba r -
                                                                        in. Hg
Person with direct responsibility  for recovered samples
Sample
No.



Final Pressure
in. Hg
Leq Af



Leq Bf



P£



Final Temperature
*F 
-------
                        NOX LABORATORY DATA  FORM
 Plant
 Date  samples received
 Aliquot factor 	
 Blank  absorbance
 Calibration  factor  (KC>
Run No. (B)
Date analyzed
Samples analyzed  by
Data reviewed  by  	
Date of review
Sample
No.



Sample
absorbance
A



Dilution
factor
F



Total mass of NO
at NO, in sample*
tn



2 K  AT,  Note:   If other than a 25 ml aliquot  ia used for analysis,  the factor
2 must be replaced by a corresponding factor.
             METHOD  7  ANALYTICAL DATA SHEET
                                 11-20

-------
               OPTIMUM WAVELENGTH  DETERMINATION1


^^ectrophotometer  No.	

 Calibrated  by 	 	
                                  Date
                                  Reviewed by
Spec tropho tome ter
setting, nm
399
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
»415
416
Absorbance
of standard
ODa






Absorbance
of blank
ODb






Actual
absorbance of
ODC






  Absorbance  of  the  200  pg  N02  standard  in  a  single  beam
  spectrophotometer.
                              a  single  beam  spectrophoto-
Absorbance of the blank in
meter.
   For  a  single  beam  spectrophotometer  -  absorbance  of  the
   standard  minus  absorbance  of  the  blank.   For  a  double
   beam spectrophotometer  - absorbance  of the  200  yg NO2
   standard  with the  blank in the  reference  cell.

 Spectrophotometer setting for maximum  actual  absorbance
 of standard (nm)
               I  » •» i i ••-*' .1 ••..,.i .  ••••^•a.

 If the maximum  actual  absorbance  occurs  at  a  spectrophotometer setting
 of 399 or 416 nm, the  spectrophotometer  must  be repaired or
 recalibrated.

      METHOD 7 Optimum  Wavelength  Determination  Data Sheet.
                            11-21

-------
    O
    z
    c
    0)
   -a
       o
       z
01
rJ

11
a
o
       4J
       HI
       en
                              01
                              «



























C K
0 O
H 4-1






















o
z o
z
X

C
0
4->
(0 IX
u o
a
X -H 11

CO O 50 <0 3
m 
01 10
Oit-i  E
u
10
01 -H 01
tl 4-1 01 O
3 c en -H
en 0) o *4-i
U 01 U V4
M W rt O
cu w
rH
Q

4J
•H T>
O 10
O 01
•H X
01

o
(0
4J
CO


























O
-O  -M
^ 3
0 01 O
o s j:
U H f
CN
01 C
C r4
H 01 1=
-i E
a-H -
g EH —

r4
O
a.
U CD
(/I &
11 3
> Z
EH





















































































































































































































































































































































































































































































































































































































































































































































































































































































































































































I


























































































































































                                                                                                                    5-1
                                                                                                                    O
(ti
-P
                                                                                                                    QJ
                                                                                                                    -H
                                                                                                                    CO
                                                                                                                    O
                                                                                                                    43
                                                                                                                    -P
                                                                                                                    Q)
                                                   11-22

-------
PLANT_
SAMPLE LOCATION.
                                 FIELD DATA CHECKS
SAMPLE RECOVERY PERSONNEL
PERSON WITH DIRECT RESPONSIBILITY  FOR RECOVERED SAMPLES.
Sample
no.
1
2
3
BLANKS
Sample
identification
no.
H2S04




so2




Date of
recovery




Liquid
level
marked




Stored
in locked
container




REMARKS
SIGNATURE OF FIELD SAMPLE TRUSTEE
                             LABORATORY  DATA CHECKS




    PERSON WITH DIRECT RESPONSIBILITY  FOR RECOVERED SAMPLES.




DATE RECOVERED SAMPLES RECEIVED	




ANALYST                                            	
Sample
no.
1
2
3
BLANKS
Sample
identification
no.
H2S04




so2




Date of
analysis




Liquid
at marked
level




Sample
identified




REMARKS
SIGNATURE OF LAB SAMPLE TRUSTEE
                   METHOD 8   Sample  recovery and integrity data.




                                   11-23

-------
    PLANT
    SAMPLE LOCATION
                                              DATE

                                       ANALYST 	
                                     1.
    NORMALITY OF BARIUM PERCHLORATE  2.

                                     3.
                                              ml Ba(C104)2
                                              ml Ba(C104)2  N(Avg
                                              ml Ba(C104J2
       SULFURIC ACID MIST  (INCLUDING SULFUR TRIOXIDE) ANALYSIS
V  ,   - Total volume of solution in which the
 so n   sulfuric acid sample is contained, ml

V  - Volume of sample aliquot, ml
 a
V  - Volume of barium perchlorate  1st titration
     titrant used for sample, ml   2nd titration
                                         Average
V   - Volume of barium perchlorate 1st titration
      titrant used for blank, ml   2nd titration
                                         Average

1st titration = Q 99 to 1 01 or
2nd titration
                                                  Run 1
                                                         Run 2
Run 3
                                     titration _ 2nd titration | ^ 0.2 ml


                        SULFUR DIOXIDE ANALYSIS
V  i  - Total volume of solution in which the
  °     sulfur dioxide sample is contained, ml
V  - Volume of sample aliquot, ml
 a
V  - Volume of barium perchlorate  1st titration
     titrant used for sample, ml   2nd titration
                                         Average

V. .  - Volume of barium perchlorate 1st titration
      titrant used for blank, ml   2nd titration
                                         Average
                                                  Run 1
                                                         Run 2
Run 3
1st titration
2nd titration

SIGNATURE OF ANALYST
              = 0.99 to 1.01 or  |1st titration - 2nd titration| ^ 0.2 ml
SIGNATURE OF REVIEWER OR SUPERVISOR
                         Method 8 analytical data sheet,
                                   11-24

-------
Plant- ;
Si-l-p-
Date •

Front Rinse

Back Rinse
Solution:
Volume: Ini
Clean Up By:
City?
Sam. Type:
Rnn No:

Front Filter Front Solu

Back Filter | ~| Back Solu
T,(=\7fai w^rk^d «
^
^
f-ial Final rg
a




Example  sample  label.
U—25    * U'S-
                       PRINTING OFFICE: 1978— 640-O13/4167 Reoion No. 4
            NOVEMBER 9

-------
TECHNICAL REPORT DATA
(Please read Instructions on the reverse before completing)
1 REPORT NO
2
4 TITLE AND SUBTITLE
Field Sampling Manual
7 AUTHOR(S)
9 PERFORMING ORG \NIZATION NAME AND ADDRESS
PEDCo Environmental, Inc.
11499 Chester Road
Cincinnati, Ohio 45246
12. SPONSORING AGENCY NAME AND ADDRESS
Division of Stationary Source Enforcemen
U.S. Environmental Protection Agency
Washington, D.C. 20460
3 RECIPIENT'S ACCESSION NO.
5 REPORT DATE
September 1978
6 PERFORMING ORGANIZATION CODE
8 PERFORMING ORGANIZATION REPORT NO.
P/N 3370-2-K
10 PROGRAM ELE MENT NO.
11 CONTRACT/GRANT NO
68-01-4147, Task
54
13 TYPE OF REPORT AND PERIOD COVERED
Final
^- 14. SPONSORING AGENCY CODE

15. SUPPLEMENTARY NOTES
DSSE Project Officer: K. Foster
16. ABSTRACT
The field sampling manual was designed to provide field en-
forcement personnel with a source sampling reference manual suitable
for field use. It contains a double-spaced, typed version of the
EPA New Source Performance Standards Test Methods 1 through 8 and
sample data sheets specifically designed for use with the revised
methods to obtain all necessary information.
17.
a DESCRIPTORS
KEY WORDS AND DOCUMENT ANALYSIS
b IDENTIFI
Air Pollution ^^e
Emission Measurements _
_ ,. . New So
Enforcement ance
1S DISTRIBUTION STATEMENT
Unlimited
19 SECURI
Un
20 SECURI
Un
ERS/OPEN ENDED TERMS c COSATI Field/Group
Manual 13B
Sampling
urce Perform- 14D
Standards
TY CLASS (ThisReport/ 21 NO. OF PAGES
classified 204 p.
TY CLASS (Thispage) 22 PRICE
classified



EPA Form 2220-1  (9-73)

-------

-------
   u
   UJ
   (3


  il
  2 1-


  oz
  2°
   M

   3
       CNJ

       q
       z
o
z
UJ
O
<
Z

9S«
I-.2 o
075£
££6   .
£.2   if
^ C 0> u) Q
g C O 8 O.
5^ £o •

ji«»?
Z°^  -S
UJ g «  H-
5 ^ J3  r-

§5 =
CC
UJ
        U

        (0
        0>
        (A
        0)
        OC
                                                   I  If
                                                   "T  »- <-'
                                                   fl)  I— «.<
                                                   >  eg .1=
                                                   2  « c
                                                   A  = I-.
                                                   «  = 3
                                                   0)  •§ *
                                                   £  « '-
                                                   c  «12
                                                   g>8 °»*
                                                   m » •£ J?
                                                   •s 2 3 *
                                                   D.A C *-
                                                   8ssg
                                                   = E £ 5
                                                   .2 3 '5 3C
                                                     --i
•o «
Si:^
                                                   3 0 3 <[>
                                                   o ^ o
                                                              i
                                                             00
O
^
CO
<
Q.
UJ
 •
O
Z
z
o

<
o
_l
CD

a

-------