5547                                                               905R80111
                         METHOD 608.2:   ANALYSIS  OF CERTAIN  ORGANOCHLORINE
                          PESTICIDES IN WASTEWATER BY GAS CHROMATOGRAPHY
          1.  SCOPE AND APPLICATION

              1.1  This method covers the determination of certain organochlorine
                   pesticides in industrial and municipal wastewater.  The following
                   parameters may be determined by this method.

                   Parameter                   STORET No.               CAS No.

                   Chlorothalonil                                      1897-45-6
                   DCPA                          39770                 1861-32-1
                   Dichloran                                             99-30-9
                   Methoxychlor                  39480                   72-43-5
                   Permethrin                                         52645-53-1

              1.2  The estimated detection limit (EDL) for each parameter is listed in
                   Table 1.  The EDL was calculated from the minimum detectable
                   response of the electron capture detector equal to 5 times the
                   detector background noise assuming a 10.0 mL final extract volume
                   of a 1 L reagent water sample and a gas chromatographic (GC)
                   injection volume of 5 uL.  The EDL for a specific wastewater  may be
                   different depending on the nature of interferences in the sample
                   matrix.

              1.3  This is a GC method applicable to the determination of the
                   compounds listed above in municipal and industrial discharges.
                   When this method is used to analyze unfamiliar samples for any or
                   all of the compounds listed above, compound identifications should
                   be supported by at least one additional qualitative technique.
                   Section 13 provides gas chromatograph/mass spectrometer (GC/MS)
                   conditions appropriate for the qualitative confirmation of compound
                   identifications.

              1.4  This method is restricted to use by or under the supervision  of
                   analysts experienced in the operation of gas chromatographs and in
                   the interpretation of chromatograms.

          2.  SUMMARY OF METHOD

              2.1  Organochlorine pesticides are removed from the sample matrix  by
                   extraction with methylene chloride.  The extract is dried,
                   exchanged into hexane, and analyzed by gas chromatography. Column
                   chromatography is used as necessary to eliminate interferences
                   which may be encountered.  Measurement of the pesticides is
                   accomplished with an electron capture detector.

              2.2  Confirmatory analysis by gas chromatography/mass spectrometry is
                   recommended (Section 13) when a new or undefined sample type  is
                   being analyzed if the concentration is adequate for such
                   determination.

                      U.S.  Environmental Protection Agency
                      Region V, Library
                      230  South Dearborn  Street
                      Chicago,  Illinois  60604

-------
f
\
        Fnvlronmenta! Protection

-------
3.  INTERFERENCES

    3.1  Solvent, reagents, glassware, and other sample processing hardware
         may yield discrete artifacts and/or elevated baselines causing
         misinterpretation of gas chromatograms.  All of these materials
         must be demonstrated to be free from interferences under the con-
         ditions of the analysis by running laboratory reagent blanks as
         described in Section 9.1.

         3.1.1  The use of high-purity reagents and solvents helps to
                minimize interference problems.  Purification of solvents by
                distillation in all-glass systems may be required.

         3.1.2  Glassware must be scrupulously cleaned (1).  Clean all
                glassware as soon as possible after use by rinsing with the
                last solvent used in it.  This should be followed by
                detergent washing with hot water and rinses with tap water
                and reagent water.  It should then be drained dry and heated
                in a muffle furnace at 400°C for 15 to 30 minutes.  Solvent
                rinses with acetone and pesticide-quality hexane may be
                substituted for the muffle furnace heating.  Volumetric ware
                should not be heated in a muffle furnace.  After drying and
                cooling, glassware should be sealed and stored in a clean
                environment to prevent any accumulation of dust or other
                contaminants.  Store the glassware inverted or capped with
                aluminum foil.

    3.2  Interferences co-extracted from the samples will vary considerably
         from source to source, depending on the diversity of the industrial
         complex or municipality being sampled.  While general cleanup
         procedures are provided as part of this method, unique samples may
       .  require additional cleanup approaches to achieve the detection
         limits listed in Table 1.

4.  SAFETY

    4.1  The toxicity or carcinogenicity of each reagent used in this method
         has not been precisely defined; however, each chemical compound
         should be treated as a potential health hazard.  From this
         viewpoint, exposure to these chemicals must be reduced to the
         lowest possible level by whatever means available.  The laboratory
         is responsible for maintaining a current awareness file of OSHA
         regulations regarding the safe handling of the chemicals specified
         in this method.  A reference file of material data handling sheets
         should also be made available to all personnel involved in the
         chemical analysis.  Additional references to laboratory safety are
         available and have been identified (2-4) for the information of the
         analyst.

5.  APPARATUS AND EQUIPMENT

    5.1  SAMPLE CONTAINERS - Narrow-mouth glass bottles, 1-liter or 1-quart
         volume, equipped with polytetrafluoroethylene (PTFE)-lined screw

-------
     caps.  Wide-mouth glass bottles, 1-quart volume, equipped with
     PTFE-lined screw caps may also be used.  Prior to use, wash bottles
     and cap liners with detergent and rinse with tap and distilled
     water.  Allow the bottles and cap liners to air dry, then muffle
     the glass bottles at 400*C for 1 hour.  After cooling, rinse the
     cap liners with hexane, seal the bottles with aluminium foil,  and
     store in a dust-free environment.

     5.1.1  Automatic sampler (optional)—Must incorporate glass sample
            containers for the collection of a minimum of 250 ml.
            Sample containers must be kept refrigerated at 4*C and
            protected from light during compositing.  If the sampler
            uses a peristaltic pump, a minimum length of compressible
            silicone rubber tubing may be used.  Before use, however,
            the compressible tubing should be thoroughly rinsed with
            methanol, followed by repeated rinsings with distilled  water
            to minimize the potential for contamination of the sample.
            An integrating flow meter is required to collect
            flow-proportional composites.

5.2  KUDERNA-DANISH (K-D) GLASSWARE

     5.2.1  Synder column—Three-ball macro (Kontes K-503000-0121 or
            equivalent).

     5.2.2  Concentrator tube—10-mL, graduated (Kontes K-570050-1025 or
            equivalent) with ground glass stopper.

     5.2.3  Evaporative flask—500-mL .(Kontes K-570001-0500 or
            equivalent).  Attach to concentrator tube with springs.

5.3  GAS CHROMATOGRAPHY SYSTEM

     5.3.1  The gas chromatograph must be equipped with a glass-lined
            injection port compatible with the detector to be used.  A
            data system is recommended for measuring peak areas.

            5.3.1.1 Column 1—180 cm long by 2 mm ID, glass, packed with
                    1.5 percent OV-17/1.95 percent OV-210 on Chromosorb
                    W-HP (100/120 mesh) or equivalent.

            5.3.1.2 Column 2—180 cm long x 2 mm ID, glass, packed  with
                    4-percent SE-30/6-percent SP-2401 on Supelcoport
                    (100/120 mesh) or equivalent.  Guidelines for the
                    use of alternate column packings are provided in
                    Section 10.3.1.

            5.3.1.3 Detector—Electron capture.  This detector has
                    proven effective in the analysis of wastewaters for

-------
                         the  parameters  listed  in  the  scope  and was  used  to
                         develop  the method  performance  statements  in  Section
                         12.   Guidelines  for the use of  alternate detectors
                         are  provided  in  Section 10.3.1.

    5.4  CHROMATOGRAPHIC COLUMN—400  mm  long x 19 mm  ID Chromaflex, equipped
         with coarse fritted  bottom plate and PTFE stopcock.  (Kontes
         K-420540-0224 or equivalent).

         CHROMATOGRAPHIC COLUMN—300  mm  long x 10 mm  ID, equipped with
         coarse fritted  bottom plate  and PFTE stopcock  (Kontes K-430540-0213
         or equivalent).

    5.5  DRYING COLUMN—Approximately 400 mm long x 20 mm ID borosilicate
         glass, equipped with coarse  fitted bottom plate.

    5.6  MISCELLANEOUS

         5.6.1  Balance—analytical,  capable of accurately weighing to the
                nearest 0.0001 g.

         5.6.2  Separatory funnel—two-liter, equipped with PTFE stopcock.

         5.6.3  Water bath—heated with concentric ring cover, capable of
                temperature control (±20C).  The  bath should be used  in  a
                hood.

         5.6.4  Standard solution storage containers—15-mL bottles with
                PTFE-lined screw caps.

         5.6.5  Boiling chips—approximately 10/40 mesh.  Heat to 400*C for
                30 minutes,  or Soxhlet extract overnight with methylene
                chloride.

6.  REAGENTS AND CONSUMABLE  MATERIALS

    6.1  REAGENTS

         6.1.1  Acetone,  hexane,  ethanol and methylene
                chloride—demonstrated to be free  of analytes.

         6.1.2  Ethyl ether—Nanograde,  redistilled in glass if necessary.
                Must be free of peroxides as indicated by EM Quant  test
                strips.  (Available from Scientific Products Co.,  Cat. No.
                P1126-8,  and other  suppliers.)   Procedures  recommended for
                removal of peroxides are provided  with the  test  strips.
                After cleanup, 20 mL ethyl  alcohol preservative  must be
                added to  each liter of ether.

         6.1.3  Florisil—PR grade  (60/100  mesh).   Purchase  activated  at
                1250°C  and store  in dark in  glass  containers with  glass
                stoppers  or  foil-lined screw caps.  Before  use,  activate

-------
                each batch overnight at 130*C in foil-covered glass
                container.

         6.1.4  Silica^gel—Activate approximately 100 grams of silica gel
                at 200°C for 16 hours in a tared 500-ml Erlenmeyer flask
                with ground glass stopper.  Allow to cool to room
                temperature, and determine the weight of activated silica
                gel.  Deactivate by adding 3 percent by weight of distilled
                water.  Restopper the flask, and shake on a wrist-action
                shaker for at least 1 hour.  Allow to equilibrate for 3 or
                more hours at room temperature.

         6.1.5  Reagent water—Reagent water is defined as a water in which
                an interferent is not observed at the method detection limit
                of each parameter of interest.

         6.1.6  Sodium hydroxide (NaOH) solution (ION)—dissolve 40 g NaOH
                in reagent water and dilute to 100 ml.

         6.1.7  Sodium sulfate—granular, anydrous.  Condition by heating at
                400°C for 4 hours in a shallow tray.

         6.1.8  Sulfuric acid (1^504) solution (1+1)—add measured
                volume of concentrated H2S04 to equal volume of reagent
                water.

    6.2  STANDARD STOCK SOLUTIONS (1.00 wg/uL)—These solutions may be
         purchased as certified solutions or prepared from pure standard
         materials using the following procedures.

         6.2.1  Prepare standard stock solutions by accurately weighing
                about 0.0100 grams of pure material.  Dissolve the material
                in hexane or other suitable solvent and dilute to volume in
                a 10-mL volumetric flask.  Larger volumes can be used at the
                convenience of the analyst.  If compound purity is certified
                at 96% or greater, the weight can be used without correction
                to calculate the concentration of the standard stock.

         6.2.2  Store standard stock solutions at 4*C in 15-mL bottles
                equipped with PTFE-lined screw-caps.  Standard stock
                solutions should be checked frequently for signs of
                degradation or evaporation, especially just prior to
                preparing calibration standards from them.

         6.2.3  Standard stock solutions must be replaced after 6 months or
                sooner, if comparison with check standards indicates a
                problem.

7.  SAMPLE COLLECTION, PRESERVATION, AND STORAGE

    7.1  Collect all samples in duplicate.  Grab samples must be collected
         in glass containers.  Conventional sampling practices (5) should be

-------
         followed, except that the bottle must not .be prewashed with sample
         before collection.

    7.2  The samples must be iced or refrigerated at 4*C from the time of
         collection until extraction.  Chemical preservatives should not be
         used in the field unless more than 24 hours will elapse before
         extraction.  If the samples will not be extracted within 48 hours
         of collection, the sample should be adjuted to a pH range of 6.0 to
         8.0 with sodium hydroxide or sulfuric acid.

    7.3  All samples must be extracted within 7 days of collection, and
         analyzed within 40 days of extraction. (6)

8.  CALIBRATION AND STANDARDIZATION

    8.1  CALIBRATION

         8.1.1  A set of at least three calibration solutions containing the
                method analytes is needed.  One calibration solution should
                contain each analyte at a concentration approaching but
                greater than the EDL (Table 1) for that compound; the other
                two solutions should contain analytes at concentrations that
                bracket the range expected in samples.  For example, if the
                detection limit for a particular analyte is 0.2 ug/L, and a
                sample expected to contain approximately 5 ug/l is analyzed,
                standard solutions should be prepared at concentrations
                representing 0.3 ug/L, 5 ug/L, and 10 yg/L of the analytes.

         8.1.2  To prepare a calibration solution, add an appropriate volume
                of a standard stock solution to a volumetric flask and
                dilute to volume with hexane.

         8.1.3  Starting with the standard of lowest concentration, analyze
                each calibration standard according to Section 10.3.2 and
                tabulate peak height or area responses versus the mass of
                analyte injected.  The results can be used to prepare a
                calibration curve for each compound.  Alternatively, if the
                ratio of response to concentration (calibration factor) is a
                constant over the working range (<10% relative standard
                deviation), linearity through the origin can be assumed and
                the average ratio or calibration factor can be used in place
                of a calibration curve.

         8.1.4  The working calibration curve or calibration factor must be
                verified on each working day by the measurement of one or
                more calibration standards.  If the response for any analyte
                varies from the predicted response by more than ±10%, the
                test must be repeated using a fresh calibration standard.
                If the results still do not agree, generate a new
                calibration curve.

-------
    8.2  FLORISIL STANDARDIZATION

         8.2.1  Florisil from different batches or sources may vary in
                absorptive capacity.  To standardize the amount of Florisil
                which may be used in the cleanup procedure (Section 10.2.2)
                use of the lauric acid value (7) is suggested.  The
                referenced procedure determines the adsorption from hexane
                solution of lauric acid (mg) per gram Florisil.  The amount
                of Florisil to be used for each column is calculated by
                dividing this factor into 110 and multiplying by 20 g.

9.  QUALITY CONTROL

    9.1  MONITORING FOR INTERFERENCES

         Analyze a laboratory reagent blank each time a set of samples is
         extracted.  A laboratory reagent blank is a one-liter aliquot of
         reagent water.  If the reagent blank contains a reportable level of
         any analyte, immediately check the entire analytical system to
         locate and correct for possible interferences and repeat the test.

    9.2  ASSESSING ACCURACY

         9.2.1  After every 10 samples, and preferably in the middle of each
                day, analyze a laboratory control standard.  Calibration
                standards may not be used for accuracy assessments and the
                laboratory control standrd may not be used for calibration
                of the analytical system.

                9.2.1.1 Laboratory Control Standard Concentrate - from stock
                        standards prepared as described in Section 6.3,
                        prepare a laboratory control standard concentrate
                        that contains each analyte of interest at a
                        concentration of 2 ng/mL in acetone or other
                        suitable solvent. (8)

                9.2.1.2 Laboratory Control Standard - using a pipet, add
                        1.00 mL of the laboratory control standard
                        concentrate to a one-liter aliquot of reagent water.

                9.2.1.3 Analyze the laboratory control standard as described
                        in Section 10.  For each analyte in the laboratory
                        control standard, calculate the percent recovery
                        (Pj) with the equation:


                                    loo s1
                        where S-j = the analytical results from the
                                   laboratory control standard, in ug/L; and
                              T-j = the known concentration of the spike,
                                   in

-------
         9.2.2  At least annually, the laboratory should participate in
                formal performance evaluation studies, where solutions of
                unknown concentrations are analyzed and the performance of
                all participants is compared.

    9.3  ASSESSING PRECISION
         9.3.1
         9.3.2
Precision assessments for this method are based upon the
analysis of field duplicates (Section 7.1).  Analyze both
sample bottles for at least 10% of all samples.  To the
extent practical, the samples for duplication should contain
reportable levels of most of the analytes.

For each analyte in each duplicate pair, calculate the
relative range (RR-j) with the equation:
              100 R,
                     RR.
         9.3.3
10. PROCEDURE
               A.

where   Ri =    the absolute difference between the
                duplicate measurements X^ and X2, in
                ug/L; and
        X-j =    the average concentration found ([Xj. +
                X2]/2), in ug/L.

Individual relative range measurements are pooled to
determine average relative range or to develop an expression
of relative range as a function of concentration.
    10.1 SAMPLE EXTRACTION

         10.1.1 Mark the water meniscus on the side of the sample bottle for
                later determination of sample volume.  Pour the entire
                sample into a 2-liter separatory funnel.  Check the pH of
                the sample with wide-range pH paper and adjust to within the
                range of 5 to 9 with sodium hydroxide or sulfuric acid.

         10.1.2 Add 60 mL of methylene chloride to the sample bottle and
                shake for 30 seconds to rinse the walls.  Transfer the
                solvent to the separatory funnel and extract the sample by
                shaking the funnel for 2 minutes with periodic venting to
                release vapor pressure.  Allow the organic layer to separate
                from the water phase for a minimum of 10 minutes.  If the
                emulsion interface between layers is more than one-third the
                volume of the solvent layer, the analyst must employ
                mechanical techniques to complete the phase separation.  The
                optimum technique depends on the sample, but may include
                stirring, filtration of the emulsion through glass wool, or
                centrifugation.  Collect the extract in a 250-mL Erlenmeyer
                flask.

-------
10.1.3 Add an additional 60-mL volume of methylene chloride to the
       sample bottle and complete the extraction procedure a second
       time, combining the extracts in the Erlenmeyer flask.

10.1.4 Perform a third extraction in the same manner.  Pour the
       combined extract through a drying column containing about 10
       cm of anhydrous sodium sulfate, and collect the extract in a
       500-mL K-D flask equipped with a 10 mL concentrator tube.
       Rinse the Erlenmeyer flask and column with 20 to 30 ml of
       methylene chloride to complete the quantitative transfer.

10.1.5 Add one or two clean boiling chips to the flask and attach a
       three-ball Snyder column.  Prewet the Snyder column by
       adding about 1 ml of methylene chloride to the top.  Place
       the K-D apparatus on a hot water bath (60 to 65*C) so that
       the concentrator tube is partially immersed in the hot water
       and the entire lower rounded surface of the flask is bathed
       in steam.  Adjust the vertical position of the apparatus and
       the water temperature as required to complete the
       concentration in 15 to 20 minutes.  At the proper rate of
       distillation, the balls of the column will actively chatter
       but the chambers will not flood.  When the apparent volume
       of liquid reaches about 3 mL, remove the K-D apparatus and
       allow it to drain and cool for at least 10 minutes.

10.1.6 Increase the temperature of the hot water bath to about 80
       to 85*C.  Momentarily remove the Snyder column, add 50 mL of
       hexane and a new boiling chip, and reattach the Snyder
       column*  Pour about 1 mL of hexane into the top of the
       Snyder column, and concentrate the solvent extract as
       before.  Elapsed time of concentration should be 5 to 10
       minutes.  When the apparent volume of liquid reaches about
       3 mL, remove the K-D apparatus, and allow it to drain at
       least 10 minutes while cooling.  Remove the Snyder column,
       rinse the flask and the lower joint into the concentrator
       tube with 1 to 2 mL of hexane, and adjust the volume to
       10 mL.  A 5-mL syringe is recommended for this operation.
       Stopper the concentrator tube, and store refrigerated if
       further processing will not be performed immediately.  If
       the extracts will be stored longer than 2 days, they should
       be transferred to Teflon-sealed screw-cap bottles.  If the
       sample extract requires no cleanup, proceed with gas
       chromatographic analysis.

10.1.7 If the sample requires cleanup, the extract obtained must be
       divided into two fractions.  One of the fractions is eluted
       through Florisil for the analysis of dicloran and DCPA.  The
       other fraction is eluted through silica gel for the analysis
       of chlorothalonil, methoxychlor, and the permethrins.

10.1.8 Determine the original sample volume by refilling the sample
       bottle to the mark and transferring the liquid to a 1,000 mL
       graduated cylinder.  Record the sample volume to the nearest
       5 mL.

-------
10.2 CLEANUP AND SEPARATION

     10.2.1 Cleanup procedures may not be necessary for a relatively
            clean sample matrix.  The cleanup procedures recommended in
            this method have been used for the analysis of various clean
            waters and municipal effluents.  The single-operator
            precision and accuracy data in Table 2 were gathered using
            the recommended cleanup procedures.   If particular
            circumstances demand the use of an alternative cleanup
            procedure, the analyst must determine the elution profile
            and demonstrate that the recovery of each compound of
            interest is no less than that recorded in Table 2.

     10.2.2 Florisil Column Cleanup.

            10.2.2.1 Add a weighed amount of Florisil, about 21 grams,
                     to a chromatographic column.  The exact weight
                     should be determined by calibration. (7)  Tap the
                     column to settle the Florisil.  Add a 1- to 2-cm
                     layer of sodium sulfate above the Florisil.  Rinse
                     the Florisil and sodium sulfate by adding 60 ml of
                     hexane to the column.  Just prior to exposure of
                     the sodium sulfate to air,  stop the draining of the
                     hexane by closing the stopcock on the column.
                     Discard the eluate.

            10.2.2.2 Quantitatively, add the fraction of extract chosen
                     for the analysis of dichloran and DCPA to the
                     column.  Drain the column into the flask, stopping
                     just prior to exposure of the sodium sulfate
                     layer.

            10.2.2.3 Elute the column with 200 mL of 6-percent ethyl
                     ether in hexane (Fraction 1) using a drip rate of
                     about 5 mL/minute.  Remove  and discard.  Perform a
                     second elution using 200 ml of 15-percent ethyl
                     ether in hexane (Fraction 2), collecting the eluant
                     in a 500 mL K-D flask equipped with a 10 mL
                     concentrator tube.

            10.2.2.4 Concentrate the eluate by standard K-D techniques
                     (Paragraph 10.1.5), substituting hexane for
                     methylene chloride, and using the water bath at
                     about 85°C.  Adjust the final volumes to 10 mL with
                     hexane.  Analyze by gas chromatography.

     10.2.3 Silica Gel Column Cleanup.

            10.2.3.1 Prepare silica gel columns  using a 300-mm by 10-mm
                     ID glass column.

-------
                     Rinse column with hexane.  Add approximately 50 mL
                     of hexane to the empty column.  Add 3.5 grams of
                     3-percent deactivated silica gel (Paragraph
                     6.1.4).  Pack by rotating slowly to release air
                     bubbles.  Top with 1.5 cm of Na2S04-  Drain
                     hexane to the top of the
            10.2.3.2 Add the fraction of extract chosen for the analysis
                     of chlorothalonil, methoxychlor, and the
                     permethrins to the column.  Open the stopcock and
                     allow it to drain to the surface of the sodium
                     sulfate.  Elute with the following solutions:

                     1st fraction — 25 ml of hexane,
                     2nd fraction — 25 mL of 6-percent MeClj in
                         hexane (volume/volume), and
                     3rd fraction — 25mL of 50-percent MeCl2 in
                         hexane.

            10.2.3.3 Collect the third fraction in a 500 ml K-0 flask
                     equipped with a 10 mL concentrator tube, and add 50
                     mL of hexane.  Concentrate on an 85°C water bath to
                     10.0 mL as described in Section 10.1.5.

     10.2.4 The elution profiles obtained in these studies are listed in
            Tables 3 and 4 for the convenience of the analyst.  The
            analyst must determine the elution profiles and demonstrate
            that the recovery of each compound of interest is no lass
            than that reported in Table 2 before the analysis of any
            samples utilizing these cleanup procedures.

     10.2.5 Proceed with gas chromatography.

10.3 GAS CHROMATOGRAPHIC ANALYSIS

     10.3.1 Recommended columns and detector for the gas chromatographic
            system are described in Section 5.3.1.  Table 1 summarizes
            the recommended operating conditions for the gas
            chromatograph.  Included in this table are estimated
            retention times and detection limits that can be achieved by
            this method.  Examples of the separations achieved by Column
            1 are shown in Figures 1 and 2.  Other packed columns,
            chromatographic conditions, or detectors may be used if data
            quality comparable to Table 2 are achieved.  Capillary
            (open-tubular) columns may also be used if the relative
            standard deviations of responses for replicate injections
            are demonstrated to be less than 6 percent and data quality
            comparable to Table 2 are achieved.

     10.3.2 Inject 2 to 5 uL of the sample extract using the
            solvent-flush technique (9).  Record the volume injected to
            the nearest 0.05 pL, the total extract volume, the fraction

-------
                of total extract utilized in each cleanup scheme and the
                resulting peak size in area or peak height units.

         10.3.3 The width of the retention time window used to make
                identifications should be based upon measurements of actual
                retention time variations of standards over the course of
                the day.  Three times the standard deviation of a retention
                time for a compound can be used to calculate a suggested
                window size; however, the experience of the analyst should
                weigh heavily in the interpretation of chromatograms.

         10.3.4 If the response for the peak exceeds the working range of
                the system, dilute the extract and reanalyze.

         10.3.5 If the measurement of the peak response is prevented by the
                presence of interferences, further cleanup is required.

11.  CALCULATIONS

    11.1 Determine the concentration (C) of individual compounds in the
         sample in pg/L with the equation:

                                   (A) (V ) (V )
                               C =       l    c
                where  A = amount of material injected, in nanograms;

                      V. = volume of extract injected, uL;

                      V. = volume of total extract, pL;

                      V  = volume of water extracted, mL;

                      V  = volume of final extract after cleanup (pL)

                      Vf = volume of extract utilized for cleanup scheme

                           (u/L)


    11.2 Report the results for the unknown samples in ug/L.  Round off the
         results of the nearest 0.1 pg/L or two significant figures.

12. METHOD PERFORMANCE

    12.1 Estimated detection limits (EDL) and associated chromatographic
         conditions are listed in Table 1(10).  The detection limits were
         calculated from the minimum detectable response of the EC detector
         equal  to 5 times the background noise, assuming a 10.0-mL final
         extract volume of a 1-liter sample and a GC injection of 5 uL.

    12.2 Single laboratory accuracy and precision studies were conducted by
         Environmental Science and Engineering, Inc. (6), using spiked
         industrial wastewater samples.  The results of these studies are
         presented in Table 2.

-------
13. GC/MS CONFIRMATION

    13.1 It is recommended that GC/MS techniques be judiciously employed to
         support qualitative identifications made with this method.  The
         mass spectrometer should be capable of scanning the mass range from
         35 AMU to a mass 50 AMU above the molecular weight of the
         compound.  The instrument must be capable of scanning the mass
         range at a rate to produce at least 5 scans per peak, but not to
         exceed 7 scans per peak utilizing a 70-V (nominal) electron energy
         in the electron impact ionization mode.  A GC to MS interface
         constructed of all-glass or glass-lined materials is recommended.
         A computer system should be interfaced to the mass spectrometer
         that allows the continuous acquisition and storage on machine-
         readable media of all  mass spectra obtained throughout the duration
         of the chromatographic program.

    13.2 Gas chromatographic columns and conditions should be selected for
         optimum separation and performance.  The conditions selected must
         be compatible with standard GC/MS operating practices.
         Chromatographic tailing factors of less than 5.0 must be achieved.
         The calculation of tailing factors is illustrated in Method 625.(11)

    13.3 At the beginning of each day that confirmatory analyses are to be
         performed, the GC/MS system must be checked to see that all DFTPP
         performance criteria are achieved.(12)

    13.4 To confirm an identification of a compound, the background
         corrected mass spectrum of the compound must be obtained from the
         sample extract and compared with a mass spectrum from a stock or
         calibration standard analyzed under the same chromatographic
         conditions.  It is recommended that at least 25 nanograms of
         material be injected into the GC/MS.  The criteria below must be
         met for qualitative confirmation.

         13.4.1 The molecular ion and other ions that are present above
                10-percent relative abundance in the mass spectrum of the
                standard must be present in the mass spectrum of the sample
                with agreement  to plus or minus 10 percent.  For example, if
                the relative abundance of an ion is 30 percent in the mass
                spectrum of the standard, the allowable limits for the
                relative abundance of that ion in the mass spectrum for the
                sample would be 20 to 40 percent.

         13.4.2 The retention time of the compound in the sample must be
                within 6 seconds of the same compound in the standard
                solution.

         13.4.3 Compounds that  have similar mass spectra can be explicitly
                identified by GC/MS only on the basis of retention time data.

    13.5 Where available, chemical ionization mass spectra may be employed
         to aid in the qualitative identification process.

-------
13.6 Should these MS procedures fail to provide satisfactory results,
     additional steps may be taken before reanalysis.  These may include
     the use of alternate packed or capillary GC columns or additional
     cleanup.

-------
REFERENCES

 1. ASTM Annual Book of Standards, Part 31, 03694, "Standard Practice for
    Preparation of Sample Containers and for Preservation," American Society
    for Testing and Materials, Philadelphia, PA, p. 679, 1980.

2.  "Carcinogens - Working with Carcinogens," Department of Health,
    Education, and Welfare, Public Health Service, Center for Disease
    Control, National Institute for Occupational Safety and Health,
    Publication No. 77-206, August 1977.

3.  "OSHA Safety and Health Standards, General Industry" (29 CFR 1910),
    Occupational Safety and Health Administration, OSHA 2206 (Revised,
    January 1976).

4.  "Safety in Academic Chemistry Laboratories," American Chemical Society
    Publication, Committee on Chemical Safety, 3rd Edition, 1979.

5.  ASTM Annual Book of Standards, Part 31, D3370, "Standard Practice for
    Sampling Water," American Society for Testing and Materials,
    Philadelphia, PA, p. 76, 1980.

6.  Test procedures for Pesticides in Wastewaters, EPA Contract Report
    #68-03-2897.  Unpublished report available from U.S. Environmental
    Protection Agency, Environmental Monitoring and Support Laboratory,
    Cincinnati, Ohio  45268.

7.  Mills, P.A., "Variation of Floricil Activity:  Simple Method for
    Measuring Adsorbent Capacity and Its Use in Standardizing Florisil
    Columns," Journal of the Association of Official Analytical Chemists,
    51, 19 (196SJ7

8.  "Handbook for Analytical Quality Control in Water and Wastewater
    Laboratories," EPA-600/4-79-019, U.S. Environmental Protection Agency,
    Environmental Monitoring and Support Laboratory - Cincinnati, Ohio
    45268, March 1979.

9.  Burke, J.A., "Gas Chromatography for Pesticide Residue Analysis; Some
    Practical Aspects," Journal of the Association of Official Analytical
    Chemists, 48, 1037 (1965).

10. "Evaluation of Ten Pesticide Methods," Contract #68-03-1760, Task  '
    No. 11,  U.S. Environmental Protection Agency, Environmental Monitoring
    and Support Laboratory, Cincinnati, Ohio  45268.

11. "Methods for Organic Chemical  Analysis of Municipal and Industrial
    Wastewater," EPA-600/4-82-057.  U.S. Environmental Protection Agency,
    Environmental Monitoring and Support Laboratory, Cincinnati, Ohio  45268.

12. Eichelberger, J.W., Harris, L.E., and Budde, W.L., Anal. Chem., 46, 1912
    (1975).

-------
                                    TABLE  1
                GAS  CHROMATOGRAPHY  OF  ORGANOCHLORINE  PESTICIDES
                         Retention Time (minutes)      Estimated Detection
Parameter
Chlorothalonil
DCPA
Dicloran
Methoxychlor
cis-Permethrin***
trans-Permethri n***
Column 1*
3.40
4.19
2.23
22.35
18.52
20.02
Column 2**
4.69
5.44
2.62
10.85
16.04
17.53
Limit (ug/L)
0.001
0.003
0.002
0.04
0.2
0.2
*   Column 1:  180 cm long by 2 mm ID, glass, packed with 1.5-percent
    OV-17/1.95-percent 0V 210 on Chromosorb W-HP (100/120 mesh) or
    equivalent; 5-percent methane/95-percent Argon carrier gas at 30 mL/min
    flow rate.  Column temperature is 200*C, detector—electron capture.

**  Column 2:  180 cm long by 2 mm ID, glass, packed with 4-percent
    SE-30/6-percent SP-2401 on Supelcoport (100/120 mesh) or equivalent;
    5-percent methane/95-percent Argon carrier gas at 60 mL/min flow rate.
    Column temperature is 200*C, detector—electron capture.
*** Column temperature is 220°C.

-------
                                    TABLE  2
                   SINGLE LABORATORY ACCURACY AND PRECISION
Parameter
Chlorothalonil

DCPA

Dicloran

Methoxychlor

cis-Permethrin

trans-Permethrin

Matrix
Type*
1
2
1
2
1
2
1
2
1
2
1
2
Spike
Range
(ug/L)
37.8
2,300
16
10,540
37.5
21,200
24.5
2,600
6.3
317
5.7
297
Number of
Replicates
7
7
7
7
7
7
7
7
7
7
7
7
Average
Percent
Recovery
84.1
94.9
77.6
89.5
98.6
90.8
102.4
102.2
99.5
77.5
78.8
88.9
Standard
Deviation
(%)
16.4
22.5
25.7
11.0
8.4
20.3
12.4
10.2
18.8
10.6
16.1
19.6
* 1 = Low-level industrial effluent.
  2 = High-level industrial effluent.

-------
                                    TABLE  3

                     ELUTION  PROFILES  FOR  FLORISIL  CLEANUP


                                               Percent Recovery by Fraction*
Parameter                                      I23~~


DCPA                                           0              99.3        0

Dicloran                                       0              96.3        0



* Eluting solvent composition for each fraction given in Section 10.2.2.3.

-------
                                    TABLE  4

                   ELUTION PROFILES FOR SILICA GEL* CLEANUP
                                    Percent Recovery by Fraction**
Parameter
Chlorothalonil
Methoxychlor
Cis-permethrin
Trans-Permethrin
1
0
0
0
0
2
0
0
0
0
3
93.8
93.8
107.2
92.5
* 3-Percent deactivated.
**Eluting solvent composition for each fraction given in Sections 10.2.3.2
  and 10.2.3.3.

-------
c
c
13
20
           6       3       10        12       14       16

                           Retention  Tine (mm.)

           GAS :::J5.0VATOG?-AM OF CHIQROTH£LO?!IL, OCrA,  DICLORUJ,  A;iO

           M£TnO
-------
U.S.  Environmental  Protection Agency
Region V, Library
230  South Dearborn  Street
      o,  Illinois  60604

-------

            e   3
           :    i
            e    oo
                 I

'  J  o

-------