5549
905R80112
                                 METHOD 643:   ANALYSIS  OF  BENTAZON
                               IN WASTEWATER BY LIQUID CHROMATOGRAPHY
          1.  SCOPE AND APPLICATION
                                             V  .f
              1.1  This method covers the determination of bentazon in municipal and
                   industrial wastewater.

                   Parameter                                            CAS No.
                   Bentazon  (Basagran)                                25057-89-0

              1.2  The estimated detection limit (EDL) for bentazon is listed in
                   Table 1.  The EOL was calculated from the minimum detectable
                   response being equal to 5 times the background noise using a 5-mL
                   final extract volume of a 1-liter sample and an injection volume of
                   100 yL.  The EDL for a specific wastewater may be different
                   depending on the nature of  interferences in the sample matrix.

              1.3  This is a high performance  liquid chromatographic (HPLC) method
                   applicable to the determination of bentazon in municipal and
                   industrial discharges.  When this method is used to analyze
                   unfamiliar samples for bentazon, compound identification should be
                   supported by at least one additional qualitative technique.  This
                   method describes analytical conditions for a second HPLC column
                   that can be used to confirm measurements made with the primary
                   column.

              1.4  This method is restricted to use by or under the supervision of
                   analysts experienced in the operation of liquid chromatographs and
                   in the interpretation of liquid chromatograms.

          2.  SUMMARY OF METHOD

              2.1  Bentazon is removed from an acidified sample matrix by extraction
                   with methylene chloride.  The extract is discarded after back
                   extraction with aqueous base.  HPLC conditions are described which
                   permit the separation and measurement of bentazon in the aqueous
                   extract.

          3.  INTERFERENCES

              3.1  Solvent, reagents, glassware, and other sample processing hardware
                   may yield discrete artifacts and/or elevated baselines causing
                   misinterpretation in liquid chromatograms.  All of these materials
                   must be demonstrated to be free from interferences under the
                   conditions of the analysis by running laboratory reagent blanks as
                   described in Section 9.1

                   3.1.1  The use of high-purity reagents and solvents helps to
                          minimize interference problems.  Purification of solvents by
                          distillation in all-glass systems may be required.
                               U.S. Environmental Protection Agency
                               Region V, Library
                               230 South  Dearborn Street
                               Chicago, Illinois  60604

-------
U.S.
Environmental Protection Agency

-------
         3.1.2  Glassware must be scrupulously cleaned (1).  Clean all
                glassware as soon as possible after use by rinsing with the
                last solvent used in it.  This should be followed by
                detergent washing with hot water and rinses with tap water
                and reagent water.  It should then be drained dry and heated
                in a muffle furnace at 400*C for 15 to 30 minutes.  Solvent
                rinses with acetone and pesticide-quality hexane may be
                substituted for the muffle furnace heating.  Volumetric ware
                should not be heated in a muffle furnace.  After drying and
                cooling, glassware should be sealed and stored in a clean
                environment to prevent any accumulation of dust or other
                contaminants.  Store the glassware inverted or capped with
                aluminum foil.

    3.2  Matrix interferences may be caused by UV-active contaminants that
         are co-extracted from the samples.  The extent of matrix
         interferences will vary considerably from source to source,
         depending upon the nature and diversity of the industrial complex
         or municipality being sampled.  Unique samples may require
         additional cleanup approaches to achieve the detection limit listed
         in Table 1.

4.  SAFETY

    4.1  The toxicity or carcinogenicity of each reagent used in this method
         has not been precisely defined; however, each chemical compound
         should be treated as a potential health hazard.  From this
         viewpoint, exposure to these chemicals must be reduced to the
         lowest possible level by whatever means available.  The laboratory
         is responsible for maintaining a current awareness file of OSHA
         regulations regarding the. safe handling of the chemicals specified
         in this method.  A reference file of material data handling sheets
         should also be made available to all personnel involved in the
         chemical analysis.  Additional references to laboratory safety are
         available and have been identified (2-4) for the information of the
         analyst.

5.  APPARATUS AND EQUIPMENT

    5.1  SAMPLE CONTAINERS - Narrow-mouth glass bottles, 1-liter or 1-quart
         volume, equipped with polytetrafluoroethylene (PTFE)-lined screw
         caps.  Wide-mouth glass bottles, 1-quart volume, equipped with
         PTFE-lined screw caps may also be used.  Prior to use, wash bottles
         and cap liners with detergent and rinse with tap and reagent
         water.  Allow the bottles and cap liners to air dry, then muffle
         the bottles at 400*C for 1 hour.  After cooling, rinse the bottle
         and cap liners-with hexane, seal the bottles, and store in a
         dust-free environment.

         5.1.1  Automatic sampler (optional)—Must incorporate glass sample
                containers for the collection of a minimum of 250 mL.
                Sample containers must be kept refrigerated at 4*C and
                protected from light during compositing.  If the sampler

-------
                uses a peristaltic pump, a minimum length of compressible
                silicone rubber tubing may be used.  Before use,  however,
                the compressible tubing should be thoroughly rinsed with
                methanol, followed by repeated rinsings with reagent water
                to minimize the potential for contamination of the sample.
                An integrating flow meter is required to collect
                flow-proportional composites.

    5.2  HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) APPARATUS  —
         Analytical system complete with liquid chromatograph and all
         required accessories including syringes, analytical columns, and
         mobile phases.  The system must be compatible with the specified
         detectors and strip-chart recorder.  A data system is recommended
         for measuring peak areas.

         5.2.1  Gradient pumping system, constant flow.

         5.2.2  Injector valve (Rheodyne 7125 or equivalent) with 100-nL
                loop.

         5.2.3  Column 1 — 250 mm by 4.6 mm ID, stainless steel, packed
                with reverse-phase Ultrasphere OOS, 10 u, or equivalent.

         5.2.4  Column 2 — 300 mm by 4.0 mm ID, packed with reverse phase
                Bondapak CIS, 10u, (Waters Associates), or equivalent.

         5.2.5  Ultraviolet detector, variable wavelength, capable of
                monitoring at 340 nm.

         5.2.6  Strip-chart recorder compatible with detector, 250-mm.  (A
                data system for measuring peak areas is recommended.)

    5.3  MISCELLANEOUS

         5.3.1  Balance—analytical, capable of accurately weighing to the
                nearest 0.0001 g.

         5.3.2  Separatory funnels—2-liter, and 250 mL, equipped with PTFE
                stopcocks.

         5.3.3  Standard solution storage containers—15-mL bottles with
                PTFE-lined screw caps.

         5.3.4  Pasteur pipets with bulbs.

6.  REAGENTS AND CONSUMABLE MATERIALS

    6.1  REAGENTS

         6.1.1  Acetone, hexane, methanol, and methylene
                chloride—Demonstrated to be free of analytes and
                interferences.

-------
     5.1.2  Reagent water—Reagent water is defined as a water in which
            an interferent is not observed at the method detection limit
            of each parameter of interest.

     6.1.3  Sodium hydroxide solution (0.1N)—Dissolve 0.4 gram of NaOH
            in reagent water and dilute to 100 ml.

     6.1.4  Sodium chloride—(ACS) Crystals.

     6.1.5  Sodium thiosulfate—(ACS) Granular.

     6.1.6  Sulfuric acid solution (1+1)—Slowly add 50 ml of H2S04
            (specific gravity 1.84) to 50 mL of reagent water.

     6.1.7  Sodium hydroxide solution (6N)—Dissolve 24 g of NaOH in
            reagent water and dilute to 100 mL.

     6.1.8  Acetate buffer solution—Dissolve 0.41 g of anhydrous sodium
            acetate (ACS) and 1.5 ml of glacial acetic acid (ACS) in 100
            mi of reagent water.

     6.1.9  Glacial acetic acid—(ACS).

     6.1.10 HPLC mobile phase buffer (pH 4.7, 0.062 M acetate)—
            Dissolve 0.87 g of anhydrous sodium acetate (ACS) and 3.0 mL
            of glacial acetic acid (ACS) in 1 L of reagent water.

6.2  STANDARD STOCK SOLUTION (1.00 wg/uL)—This solution may be
     purchased as a certified solution or prepared from a pure standard
     material using the following procedures.

     6.2.1  Prepare the stock standard solution by accurately weighing
            about 0.0100 gram of pure material.  Dissolve the material
            in pesticide quality methanol, dilute to volume in a 10-mL
            volumetric flask.  Larger volumes can be used at the
            convenience of the analyst.  When compound purity is
            certified at 96 percent or greater, the weight can be used
            without correction to calculate the concentration of the
            stock standard.  Commercially prepared stock standards can
            be used at any concentration if they are certified by the
            manufacturer or by an independent source.

     6.2.2  Transfer the stock standards in TeflonR-sealed screw-cap
            bottles.  Store at 4*C and protect from light.  Stock
            standards should be checked frequently for signs of
            degradation or evaporation, especially just prior to
            preparing calibration standards from them.

     6.2.3  Stock standards must be replaced after 6 months, or when
            comparison with quality control check samples indicates a
            problem.

-------
7.  SAMPLE COLLECTION, PRESERVATION; AND STORAGE

    7.1  Collect all samples in duplicate.  Grab samples must be collected
         in glass containers.  Conventional sampling practices (5) should be
         followed, except that the bottle must not be prewashed with sample
         before collection.

    7.2  The samples must be iced or refrigerated at 4*C from the time of
         collection until extraction.  If the samples will not be extracted
         within 48 hours of collection, the sample should be adjusted to a
         pH range of 6.0 to 8.0 with sodium hydroxide or sulfuric acid and
         35 mg of sodium thiosulfate per ppm of free chlorine per liter must
         be added.

    7.3  All samples must be extracted within 7 days and completely analyzed
         within 30 days of extraction. (6)

8.  CALIBRATION

    8.1  Establish liquid chromatographic operating parameters equivalent to
         those indicated in Table 1.

    8.2. Prepare calibration standards at a minimum of three concentration
         levels of bentazon by adding volumes of the stock standard to a
         volumetric flask and diluting to volume with HPLC mobile phase
         (35-percent methanol in HPLC mobile phase buffer or
         40-percentmethanol in HPLC mobile phase buffer).  One of the
         external standards should be at a concentration near, but greater
         than, the EDL, and the other concentrations should correspond to
         the expected range of concentrations found in real samples or
         should define the working range of the detector.

    8.3  Using injections of 100 uL of each calibration standard, tabulate
         peak height or area responses against the mass injected.  The
         results are used to prepare a calibration curve for the analytes.
         Alternatively, if the ratio of response to amount injected
         (calibration factor) is a constant over the working range (<10
         percent relative standard deviation, RSD), linearity of the
         calibration curve can be assumed and the average ratio or
         calibration factor can be used in place of a calibration curve.

    8.4  The working calibration curve or calibration factor must be
         verified on each working day by the measurement of one or more
         calibration standards.  If the response for bentazon varies from
         the predicted response by ±10 percent, the test must be repeated
         using a fresh calibration standard.  Alternatively, a new
         calibration curve or factor must be prepared.

    8.5  Before using any cleanup procedure, the analyst must process a
         series of calibration standards through the procedure to validate
         elution pattern and the absence of interferences from the reagents.

-------
9.  QUALITY CONTROL

    9.1  MONITORING FOR INTERFERENCES

         Analyze a laboratory reagent blank each time a set of samples is
         extracted.  A laboratory reagent blank is an aliquot of reagent
         water.  If the reagent blank contains a reportable level of
         bentazon, immediately check the entire analytical system to locate
         and correct for possible interferences and repeat the test.

    9.2  ASSESSING ACCURACY

         9.2.1  After every 10 samples, and preferably in the middle of each
                day, analyze a laboratory control standard.  Calibration
                standards may not be used for accuracy assessments and the
                laboratory control standrd may not be used for calibration
                of the analytical system.

                9.2.1.1 Laboratory Control Standard Concentrate  (100 ug/L)—
                        From stock standards prepared as described in
                        Section 6.2, prepare a laboratory control standard
                        concentrate that contains bentazon at a
                        concentration of 2 ug/uL in methanol.(7)

                9.2.1.2 Laboratory Control Standard — Using a pipet or
                        micro liter syringe, add 50.0 uL of the laboratory
                        control standard concentrate to a 1.0 L  aliquot of
                        reagent water contained in a 1-L volumetric flask.

                9.2.1.3 Analyze the laboratory control standard  as described
                        in Section 10.  Calculate the percent recovery
                        (Pi) with the equation:
                             P. ,  .     Si
                                       Ti

                        where S-j - the analytical results from the
                                   laboratory control standard, in ug/L; and
                              T-j m the known concentration of the spike,
                                   in ug/L.

         9.2.2  At least annually, the laboratory should participate in
                formal performance evaluation studies, where solutions of
                unknown concentrations are analyzed and the performance of
                all participants is compared.(7)

    9.3  ASSESSING PRECISION

         9.3.1  Precision assessments for this method are based upon the
                analysis of field duplicates (Sect. 7.1).  Analyze both
                samples for at least 10% of all samples.  To the extent

-------
              practical, the samples for duplication should contain
              reportable levels of bentazon.

         9.3.2  Calculate the relative range (7) (RR-j) with the equation:
                              100 R,
                     RR
                       i
                               Xi
                where   R-j -      the absolute difference between the
                                  duplicate measurements X^ and Xg, in
                                  ug/L; and
                        XT m      the average concentration found ([X^ +
                                  X2]/2), in yg/L.

         9.3.3  Individual relative range measurements are pooled to
                determine average relative range or to develop an expression
                of relative range as a function of concentration.

10. PROCEDURE

    10.1 SAMPLE EXTRACTION

         10.1.1 Mark the water meniscus on the side of the sample bottle for
                later determination of sample volume.  Pour the entire
                sample (approximately one liter) into a 2-liter separatory
                funnel.  Add 35 mg/L of sodium thiosulfate per ppm of free
                chlorine.  Check the pH of the sample with wide-range pH
                paper and adjust to within the range of 2.5 to 3.5 with
                sulfuric acid.  Add 200 grams of sodium chloride and mix to
                dissolve.

         10.1.2 Add 60 ml of methylene chloride to the sample bottle and
                shake for 30 seconds to rinse the walls.  Transfer the
                solvent to the separatory funnel and extract the sample by
                shaking the funnel for 2 minutes with periodic venting to
                release vapor pressure.  Allow the organic layer to separate
                from the water phase for a minimum of 10 minutes.  If the
                emulsion interface between layers is more than one-third the
                volume of the solvent layer, the analyst must employ
                mechanical techniques to complete the phase separation.  The
                optimum technique depends on the sample, but may include
                stirring, filtration of the emulsion through glass wool, or
                centrifugation.  Collect the extract in a 250-mL separatory
                funnel.

         10.1.3 Add an additional 60-mL volume of methylene chloride to the
                sample bottle and complete the extraction procedure a second
                time, combining the extracts in the 250-mL separatory funnel.

         10.1.4 Perform a third extraction in the same manner.  Add 2 mL of
                0.1 M NaOH in reagent water to the 250-mL separatory funnel,
                and extract by shaking the funnel for 2 minutes with
                periodic venting to release vapor pressure.  Allow the

-------
            organic layer to separate from the water phase for a minimum
            of 10 minutes.  Drain the methylene chloride into a 250-mL
            Erlenmeyer flask.  Transfer the aqueous layer with a Pasteur
            pipet to a 5-mL volumetric flask.

     10.1.5 Add the methylene chloride back to the 250-mL separatory
            funnel, and extract with an additional 2 ml of 0.1 M NaOH.
            Combine the extracts in the 5-mL volumetric flask.

     10.1.6 Add two drops of glacial acetic acid to the volumetric
            flask, and dilute to volume with acetate buffer solution
            (Section 6.1.7).

     10.1.7 Determine the original sample volume by refilling the sample
            bottle to the mark and transferring the liquid to a 1,000-mL
            graduated cylinder.  Record the sample volume to the nearest
            5 ml.

10.2 CLEANUP AND SEPARATION

     10.2.1 The cleanup procedure recommended in this method involves
            the back extraction of a methylene chloride extract with
            aqueous base, and has been used for the analysis of various
            clean waters and industrial effluents.  If additional
            cleanup is required, a one liter sample is adjusted to pH 12
            with 6N NaOH and extracted with three, 60 mL aliquots of
            methylene chloride in a 2L seporatary funnel.  The methylene
            chloride extracts are discarded and the aqueous sample
            adjusted to pH range of 2.5 to 3.5 with 1:1 sulphuric acid
            solution for re-extraction as per sect. 10.1.1.  If
            additional cleanup is required, or if particular
            circumstances demand the use of an alternate cleanup
            procedure, the analyst must determine the elution profile
            and demonstrate that the recovery for each compound of
            interest is no less than 85 percent.

10.3 LIQUID CHROMATOGRAPHY ANALYSIS

     10.3.1 Table 1 summarizes the recommended operating conditions for
            the liquid chromatograph.  Included in this table are the
            estimated retention times and estimated detection limit that
            can be achieved by this method.  An example of the
            separation achieved by Column 2 is shown in Figure 1.  Other
            columns, chromatographic conditions, or detectors may be
            used if data quality comparable to table 2 is achieved.

     10.3.2 Calibrate the system daily as described in Section 8.

10.4 Inject 100 \il of the sample extract.  Monitor the column eluent at
     340 nm.  Record the resulting peak size in area or peak height
     units.

-------
    10.5 The retention time window used to make identifications should be
         based  upon measurements of actual retention time variations of
         standards  over the course of a day.  Three times the standard
         deviation  of a retention time for a compound can be used to
         calculate  a suggested window size; however, the experience of the
         analyst  should weigh heavily in the interpretation of chromatograms.

    10.6 If the response for the peak exceeds the working range of the
         system,  dilute the sample with mobile phase and reanalyze.

    10.7 If the measurement of the peak response is prevented by the
         presence of interferences, additional cleanup is required.

11. CALCULATIONS

    11.1 Determine  the concentration of bentazon in the sample.

         11.1.1 Calculate the amount of bentazon injected from the peak
               response from the calibration curve.  The concentration in
               the sample can be calculated from the equation:
                                            (A) (Vt)
                   Concentration, ug/L »    (V ) (V )	

               where:
               A - Amount of bentazon injected (nanograms),
               V^  « Volume of extract injected (uL),
               V^  « Volume of total extract (uL), and
               Vs  * Volume of water extracted (mL).

    11.2 Report results in vg/l without correction for recovery data.  When
         duplicate  and spiked samples are analyzed, report all data obtained
         with the sample results.

12.  METHOD PERFORMANCE

    12.1 The EDL  and associated chromatographic conditions for bentazon are
         listed in  Table 1(8).  The EDL is defined as the minimum response
         being equal to 5 times the background noise, assuming a 5-mL final
         extract  volume of a 1-liter sample and an HPLC injection volume of
         100 UL.

    12.2 Single operator accuracy and precision studies were conducted by
         Environmental Science and Engineering, Incorporated (6), in the
         designated matrices.  The results of these studies are presented in
         Table  2.

-------
                                                                                          I
REFERENCES                         "                                                   "   ;
  	                                                                                     t
                                                                                          f
    1.   ASTM Annual Book of Standards, Part 31, 03694, "Standara Practice                |
         for Preparation of Sample Containers and for Preservation,"
         American Society for Testing and Materials, Philadelphia, PA, p.
         679, 1980.

    2.   "Carcinogens - Working with Carcinogens," Department of Health,
         Education, and Welfare, Public Health Service, Center for Disease
         Control, National Institute for Occupational Safety and Health,
         Publication No. 77-206, Aug. 1977.

    3.   "OSHA Safety and Health Standards, General  Industry" (29 CFR 1910),
         Occupational Safety and Health Administration, OSHA 2206 (Revised,
         January 1976).

    4.   "Safety in Academic Chemistry Laboratories," American Chemical
         Society Publication, Committee on Chemical  Safety, 3rd Edition,
         1979.

    5.   ASTM Annual Book of Standards, Part 31, D3370, "Standard Practice
         for Sampling Water," American Society for Testing and Materials,
         Philadelphia, PA, p. 76, 1980.

    6.   Test procedures for Pesticides in Wastewaters, EPA Contract Report
         No. 68-03-2897 (In Preparation).  Unpublished report available from
         U.S. Environmental Protection Agency, Environmental Monitoring and
         Support Laboratory, Cincinnati, Ohio 45268.

    7.   "Handbook for Analytical Quality Control in Water and Wastewater
         Laboratories," EPA-600/4-79-019, U.S. Environmental Protection
         Agency, Environmental Monitoring and Support Laboratory -
         Cincinnati, Ohio 45268, March 1979.

    8.   "Evaluation of Ten Pesticide Methods," U.S: Environmental
         Protection Agency Contract No. 68-03-1760,  Task No. 11, U.S.
         Environmenal Monitoring and Support Laboratory, Cincinnati, Ohio
         45268.

-------
              TABLE Z.  SINGLE LABORATORY ACCURACY AND PRECISION
Parameter
Bentazon

Matrix
Type*
1
2
Range
ug/L
125
20,400
No. of
Replicates
7
7
Average
Percent
Recovery
85.1
88.4
Standard
Deviation
(%)
4.8
8.4
*1 » 50-percent industrial effluent + 50-percent POTW effluent.
 2 » 100-percent industrial effluent.

-------
DUE
    U.S. Environmental Protection Agency
    Region V,  Library
    230 South Dearborn Street
    Chicago, Illinois  60604

-------
         u
         4)
        J
        Q     1     2     3     4     5
                 Retention Time (Min.)

FIGURE 1.  HPLC OF BENTAZON COLUMN 2

-------