R
  EPA-540A
  94-094
          United States         Office of         Publication 9240.1 -12
          Environmental Protection     Solid Waste and      EPA/540/R/94/094
          Agency           Emergency Response    ^B95-963515
                                  December 1994
          Superfund
&EFA     USEPA CONTRACT
           LABORATORY PROGRAM

           STATEMENT OF WORK
           FOR INORGANICS ANALYSIS

           MULTI-MEDIA,
           MULTI-CONCENTRATION

           ILMO1.0

-------
                                     9240.1-12
                                     PB95-963515
                                     EPA540/R-94/094
USEPA CONTRACT LABORATORY PROGRAM
        STATEMENT OF WORK
               FOR
       INORGANICS ANALYSIS
           Multi-Media
       Multi- Concentration

     Document  Number ILM01.0

-------
                                                                ATTACHMENT A
                              STATEMENT OF WORK

                              TABLE OF CONTENTS
EXHIBIT A:   SUMMARY OF REQUIREMENTS
EXHIBIT B:   REPORTING AND DELIVERABLES  REQUIREMENTS
EXHIBIT C:   INORGANIC TARGET ANALYTE  LIST
EXHIBIT D:   ANALYTICAL METHODS
EXHIBIT E:    QUALITY ASSURANCE/QUALITY CONTROL REQUIREMENTS
EXHIBIT F:    CHAIN-OF-CUSTODY, DOCUMENT  CONTROL AND STANDARD OPERATING
             PROCEDURES
EXHIBIT G:    GLOSSARY OF TERMS
EXHIBIT H:   DATA DICTIONARY AND FORMAT FOR DATA DELIVERABLES IN COMPUTER -
             READABLE FORMAT
                                                                       ILM01.0

-------
       EXHIBIT A
NUMMARY OF REQUIREMENTS
                                           ILM01.0

-------
                            GENERAL REQUIREMENTS
The Contractor shall employ procedures specified in this Statement of Work
(SOW) in the preparation and analysis of aqueous (water) and solid
(soil/sediment) samples for the presence and quantitation of 23 indicated
elements and cyanide.

The Contractor shall use proven instruments and techniques to identify and
measure the elements and inorganic species presented in the Target Analyte
List (Exhibit C).  The Contractor shall perform sample preparation and
analysis procedures as prescribed in Exhibit D, meeting specified sample
preservation and holding time requirements.

The Contractor shall adhere to the quality assurance/quality control
protocol specified in Exhibit E for all samples analyzed under this
contract.

Following sample analysis, the Contractor shall perform data reduction and
shall report analytical activities, sample data, and quality control
documentation as designated in Exhibit B.

Exhibit F contains chain-of-custody and document control requirements which
the Contractor must follow in processing samples and specifies requirements
for written laboratory standard operating procedures.

To ensure proper understanding of language utilized in this contract,
Exhibit G contains a glossary of terms.  when a term is used in the text
without explanation, the glossary meaning shall be applicable.   Glossary
definitions do not replace or take precedence over specific information
included in the SOW text.

The samples to be analyzed by the Contractor are from known or suspected
hazardous waste sites and, potentially, may contain hazardous inorganic
and/or organic materials at high concentration levels.  The Contractor
should be aware of the potential hazards associated with the handling and
analyses of these samples.  It is the Contractor's responsibility to take
all necessary measures to ensure the health and safety of its employees.

In addition, the Contractor must be aware of the importance of maintaining
the integrity of the data generated under the contracts as it is used to
make major decisions regarding public health and environmental welfare.
The data may also be used in litigation against Potentially Responsible
Parties in the enforcement of Superfund legislation.

Prior to accepting any samples from the Agency, the Contractor shall have,
in house,  the appropriate standards for all target analytes listed in
Exhibit C.
                                    A-l                                ILM01.0

-------
A.   FOR EACH SAMPLE,  THE CONTRACTOR SHALL PERFORM THE'FOLLOWING TASKS:

     Task I:      Receive and Prepare Hazardous  Waste  Samples.

     1.    The  Contractor shall  receive  and handle  samples under  the  chain-
          of-custody and sample documentation procedures described in
          Exhibit  F.  A  sample consists of all  components, perhaps more  than
          one  phase, contained inside appropriate  receptacles.   More than
          one  container  may  be used  for a  single sample; individual
          containers may contain preservatives  for  different analysis
          portions.  Containers may  be glass or plastic.

     2.    The  Contractor shall provide the  required analytical expertise and
          instrumentation for analyses of  Target Analyte List (TAL) elements
          and  cyanide equal  to or lower than the detection limits specified
          in Exhibit C.   In  Exhibit  D, EPA  provides the Contractor with  the
          specific sample  preparation techniques for water and soil/sediment
          samples and the  analytical procedures which must be used.  A
          schematic flow chart depicting the complete low level-medium level
          inorganics analytical scheme is  presented in Section I of Exhibit
          D.

     3.    The  Contractor  shall prepare and  analyze samples within the
          maximum holding  time specified in Section II of Exhibit D even if
          these times are  less than  the maximum data submission  time allowed
          in this contract.

     4.    The  Contractor  is  advised  that the samples received under this
          contract are usually from known or suspected hazardous waste sites
          and  may contain  high (greater than 15%) levels of organic and
          inorganic materials of a potentially hazardous nature and of
          unknown structure  and concentration,  and should be handled
          throughout the  analysis with appropriate caution.   It is the
          Contractor's responsibility to take all necessary measures to
          ensure laboratory  safety.

     Task  II:    Analyze Samples for Identity and Quantitation of Specific
                Inorganic Constituents.

     1.    For  each sample  received,  the Contractor may be required to
         perform the analyses described in paragraphs 2.,  3.  and 4.,
          following.  The  documentation that accompanies the sample(s)  to
          the  Contractor  facility shall indicate specific analytical
          requirements for that sample or set of samples.

     2.    Exhibit D specifies the analytical procedures that must be used.
          Exhibit D contains instructions and references for preparation of
          samples containing low-to-medium  concentrations of inorganics for
          ICP  analysis;  flame, graphite furnace and cold vapor AA analysis;
                                   A-2                                ILM01.0

-------
      and cyanide  analysis.   The  identification  and  quantitation of
      analytes  other  than  cyanide  shall be accomplished using  the  ICP or
      AA  methods specified in Exhibit  D, whichever method will achieve
      the Contract Required Detection  Limit  (CRDL) in Exhibit  C. Cyanide
      shall be  analyzed by the  individual procedures specified in
      Exhibit D.

 3.    All samples  must initially be run undiluted (i.e., final product
      of  sample preparation procedure).  When an analyte concentration
      exceeds the  calibrated or linear range, appropriate dilution (but
      not below the CRDL)  and reanalysis of  the prepared sample is
      required, as specified in Exhibit D.

 4.    For the purpose of this contract, a full sample analysis is
      defined as analysis  for all  of the target constituents identified
      in  Exhibit C in accordance with the methods in Exhibit D and
      performance of related QA/QC as specified in Exhibit E.  Duplicate
      sample,  laboratory control sample, and spike sample analyses shall
      each be considered a  separate full sample analysis   All other
     QA/QC requirements are considered an inherent part of this
      contract Statement of Work and are included in the contract sample
     unit price.

 Task  III:   Perform Required Quality Assurance/Quality Control
            Procedures

 1.   All specific QA/QC procedures prescribed in Exhibit E shall be
     strictly adhered to by the Contractor.   Records documenting the
     use of the protocol shall be maintained in accordance with the
     document control procedures prescribed in Exhibit F,  and shall be
     reported in accordance with Exhibit B requirements.

 2.   The Contractor shall establish and use on a continuing basis  QA/QC
     procedures including the daily or (as required) more frequent use
     of standard reference solutions from EPA,  the National Institute
    , of Standards  and Technology or secondary standards traceable
     thereto,  where available at appropriate concentrations (i.e.,
     standard solutions  designed to ensure that operating parameters of
     equipment and procedures, from sample receipt through
     identification and  quantitation,  produce reliable data).   Exhibit
     E specifies  the QA/QC procedures  required.

 3.   The Contractor shall establish a  Quality Assurance Plan (QAP) with
     the objective of providing sound  analytical chemical  measurements.
     This program  shall  incorporate the quality control procedures,  any
     necessary corrective action,  and  all documentation required during
     data collection as  well as the quality assessment measures
     performed by  management to ensure acceptable  data production.

4.   Additional quality  assurance and  quality control  shall be required
     in the  form of Performance Evaluation Samples  submitted by  EPA for
     Contractor analysis,  and in the form of verification  of instrument
     parameters, as  described in Exhibit E.
                               A-3                                 ILM01.0

-------
     5.   Laboratory Control Sample  (LCS)  -  This  standard  solution is
          designed Co assure that  the  operating parameters  of  the  analytical
          instrumentation and analytical  procedures  from sample  receipt
          through identification and quantitation produce  reliable data.
          The Contractor must analyze  the LCS  concurrently  with  the analysis
          of the  samples in the  SDG.

B.   EPA has provided to the Contractor formats  for  the reporting  of data
     (Exhibits B  and H).  The Contractor  shall be responsible  for  completing
     and returning analysis data sheets and submitting computer-readable
     data on diskette in the format  specified  in this  SOW  and  within the
     time specified in the Contract  Performance/Delivery Schedule.

     1.   Use of  formats other than  those  designated by EPA will be deemed
          as noncompliant.   Such data  are  unacceptable.  Resubmission in the'
          specified format at no additional  cost  to  the government will be
          required.

     2.   Computer generated forms may be  submitted  in  the  hardcopy data
          package(s)  provide^ that the forms are  in  EXACT EP«.  FORMAT.   This
          means that  the order of data elements is the  same as on  each  EPA
          required form,  including form numbers and  titles, page numbers and
          header  information, columns  and  lines.

     3.   The data reported by the Contractor  on  the hardcopy  data forms and
          the associated computer-readable data submitted by the Contractor
          must contain identical information.  If during government
          inspection  discrepancies are found,  the Contractor shall be
          required to resubmit either  or  both  sets of data  at  no additional
          cost to the government.  The resubmitted diskette must contain all
          of the  initially correct information previously  submitted for all
          samples including the  Laboratory Control Sample,  standards,  and
          blanks  in the  SDG in addition to the corrections  replacing the
          variables which were incomplete  or incorrect  according to the
          requirements in the SOW.

J.   The Contractor shall provide  analytical equipment  and technical
     expertise for this  contract as  specified  following:

     1.   Inductively coupled plasma (ICP) emission  spectrometer with the
          capability  to  analyze  metals sequentially  or  simultaneously.

     2.   Atomic  absorption (AA) spectrometer  equipped  with graphite
          furnace,  flame,  and cold vapor  AA  (or a specific  mercury analyzer)
          analysis capabilities.

     3.   Analytical  equipment/apparatus  for analysis of cyanide as
          described in Exhibit D.

D.   The minimum  functional requirements  necessary to  meet, the terms and
     conditions of this  contract are listed in items 1-7 below.  The
     Contractor shall designate  and  utilize  qualified  key  personnel to
     perform these functions. The EPA reserves  the  right  to  review


                                    A-4                                ILM01.0

-------
     personnel qualifications and experience.  See Section III, Detailed
     Technical & Management Requirements.

     1.   Laboratory Supervisor
     2.   Quality Assurance Officer
     3.   Systems Manager
     4.   Programmer Analyst
     5.   ICP Spectroscopist
     6.   ICP Operator
     7.   Atomic Absorption (AA)  Operator
     8.   Inorganic Sample Preparation Specialist
     9.   Classical Techniques (Cyanide)  Analyst
     10.   Inorganic Chemist (Backup)

E.   The  Contractor shall respond in a timely manner (within 7 days of the
     originator's request) to requests from data recipients for additional
     information or explanations  that result from  the Government's
     inspection activities.

F.   The  Contractor is required to retain unused sample volume and used
     sample containers for a period of 60 days after data submission.   From
     time of receipt until analysis,  the  Contractor shall maintain
     soil/sediment samples stored at 4°C  (+2°C).

G.   The  Contractor shall adhere  to chain-of-custody and document control
     procedures described in Exhibit F.  Documentation, as described
     therein, shall be required to show that all procedures are being
     strictly followed.   This documentation shall  be reported in the
     Complete SDG File (see Exhibit B).

H.   Sample shipments to the Contractor's facility will be scheduled and
     coordinated by the  EPA CLP Sample Management  Office (SMO),  acting on
     behalf of the Administrative Project Officer.   The Contractor shall
     communicate with SMO personnel by telephone as necessary throughout the
     proc'ess of sample scheduling, shipment,  analysis and data reporting, to
     ensure that samples are properly processed.

     If there are problems with the samples (e.g.,  mixed media,  containers
     broken or leaking)  or sample documentation/paperwork (e.g.,  Traffic
     Reports not with shipment,  or sample and Traffic Report numbers do not
     correspond) the Contractor shall immediately  contact SMO for
     resolution. The Contractor shall immediately  notify SMO regarding any
     problems and laboratory conditions that affect the timeliness of
     analyses and data reporting.  In particular,  the Contractor shall
     notify SMO personnel in advance regarding sample data that will be
     delivered late and shall specify the estimated delivery date.

I.   Sample analyses will be scheduled by groups of samples,  each defined as
     a  Case and identified by a unique EPA Case number assigned by SMO. A
     Case signifies a group of samples collected at one site or geographical
     area over a finite  time period,  and  will include one or more field
     samples with associated blanks.   Samples may  be shipped to the

                                   A-5                                ILM01.0

-------
      Contractor  in a. single  shipment  or multiple shipments  over  a.  period  of
      time,  depending on  the  size  of the Case. A Case consists of one  or more
      Sample Delivery Group(s).  A Sample  Delivery Group  (SDG) is defined  by
      the  following,  whichever  is  most frequent:

          o   each Case of field samples received,  OR

          o   each 20 field samples within a Case,  OR

          o   each 14 calendar day period  during which  field samples  in a
              Case are received (said  period beginning  with  the  receipt of
              the first sample in  the  Sample Delivery Group).

      Samples may  be  assigned to Sample Delivery Groups by matrix (i.e., all
      soils  in one  SDG, all waters  in  another), at the discretion of the
      laboratory.   Such assignment  must be made at the time  the samples are
      received, and may not be made retroactively.

     Data for all  samples in a Sample Delivery Group must be submitted
     together (in one package) in  the order specified in Exhibit B.  The
     Sample Delivery Group number  is  the EPA sample number of the first
     sample received in the SDG.  When several samples  are received together
     in the first SDG shipment, the SDG number is  the lowest sample number
     (considering both alpha and numeric designations)  in the first group of
     samples received under the SDG.  The SDG number is reported on all data
     reporting forms. The SDG Receipt Date is the  day that the last sample
     in the SDG is received.

     The Contractor is responsible for identifying each Sample Delivery
     Group as samples are received, through proper sample documentation (see
     Exhibit B)  and communication with SMO personnel.

J.   Each sample received by the Contractor will be labeled with an EPA
     sample number, and accompanied by a Traffic Report form bearing the
     sample number and descriptive information regarding the sample.   EPA
     field sample numbers are six digits in length.   If the Contractor
     receives a sample number of any other length,  contact SMO immediately.
     The'Contractor shall complete and sign the Traffic Report,  recording
     the date of sample receipt and sample condition on receipt  for each
     sample container.  The Contractor must also follow the instructions
     given on the Traffic Report in choosing the QC samples when such
     information is provided.

     The Contractor shall submit signed copies of  Traffic Reports for all
     samples in a Sample Delivery Group to SMO within 3 calendar days
     following receipt of the last sample  in the Sample Delivery Group.
     Traffic Reports shall be submitted in Sample  Delivery Group sets (i.e.,
     all Traffic  Reports for a Sample  Delivery Group shall be clipped
     together)  with an SDG Cover Sheet containing  information regarding the
     Sample Delivery Group,  as specified in Exhibit B.

K.   EPA Case numbers (including SDG numbers) and  EPA sample numbers shall
     be used by the Contractor in identifying samples received under this
     contract both verbally and in reports/correspondence.
                                    A-6                                 ILM01.0

-------
L.   Samples will routinely be shipped directly to the Contractor through a
     delivery service.  The Contractor shall be available to receive sample
     shipments at any time the delivery service is operating, including
     Saturdays and holidays.   As necessary, the Contractor shall be
     responsible for any handling or processing required for the receipt of
     sample shipments, including pick-up of samples at the nearest servicing
     airport,  bus station or  other carrier service within the Contractor's
     geographical area.

M.   The Contractor shall accept all samples scheduled by SMO,  provided that
     the total number of samples received in any calendar month does not
     exceed the monthly  limitation expressed in the contract.   Should the
     Contractor elect to accept additional samples,  the Contractor shall
     remain bound by all contract requirements  for analysis  of  those samples
     accepted.
                                  A-7                                 ILM01.0

-------
                           PERSONNEL REQUIREMENTS


I.    TECHNICAL CAPABILITY

     A.    Technical  Supervisory Personnel

          1.   Inorganics  Laboratory Supervisor

              a.   Responsible for all technical efforts of the Inorganics
                   Laboratory to meet all terms and conditions of the EPA
                   contract.

              b.   Qualifications

                   (1)  Education:

                        Minimum of Bachelor's degree in chemistry or any
                        scientific/engineering discipline.

                   (2)  Experience:

                        Minimum of three years of laboratory experience,
                        including at least one year in a supervisory
                        position.

          2.   Quality Assurance Officer

              a.   Responsible for overseeing the quality assurance aspects
                   of the data and reporting directly to upper management
                   to meet all terms and conditions of the EPA contract.

              b.   Qualifications:

                   (1)  Education:

                        Minimum of  Bachelor's degree in chemistry or any
                        scientific/engineering discipline.

                   (2)  Experience:

                        Minimum of three years of laboratory experience,
                        including at least one year of applied experience
                        with QA principles and practices in an analytical
                        laboratory.

          3.   Systems Manager

              a.   Responsible for the management and quality control of
                   all computing systems (hardware, software, documentation
                   and procedures), generating, updating, and quality
                   controlling automated deliverables to meet all terms and
                   conditions of the EPA contract.
                                   A-8                                 ILM01.0

-------
          b.   Qualifications:

               (1)  Education:

                   Minimum of Bachelor's degree with four or more
                   intermediate courses in programming, information
                   management, database management systems, or systems
                   requirements analysis.

               (2)  Experience:

                   Minimum of three years  experience in data or
                   systems management or programming including one
                   year experience with the software being utilized
                   for data management and generation of deliverables.

     4.    Programmer Analyst

          a.   Responsible for the installation, operation and
              maintenance of software and programs, generating,
              updating and quality controlling analytical databases
              and automated deliverables to meet all terms and
              conditions of the EPA contract.

          b.   Qualifications:

               (1)  Education:

                   Minimum of Bachelor's degree with four or more
                   intermediate courses in programming, information
                   management, information systems, or systems
                   requirements analysis.

               (2)  Experience:

                   Minimum of two years experience in systems or
                   applications programming including one year of
                   experience with the software being utilized for
                   data management and generation of deliverables.

B.    Technical  Staff

     1.    ICP Spectroscopist Qualifications

          a.   Education:

              Minimum of Bachelor's degree in chemistry or any
              scientific/engineering discipline.

              Specialized training in ICP Spectroscopy.
                              A-9                                 ILM01.0

-------
     b.   Experience:

          Minimum of two years of applied experience with ICP
          analysis of environmental samples.

 2.   ICP Operator Qualifications

          Minimum of Bachelor's degree in chemistry or any
          scientific/engineering discipline with one year of
          experience in operating and maintaining ICP
          instrumentation, or, in lieu of the educational
          requirement,  three additional years of experience in
          operating and maintaining ICP instrumentation.

 3.   Atomic Absorption (AA) Operator Qualifications

          Minimum of Bachelor's degree in chemistry or any
          scientific/engineering discipline with one year of
          experience in operating and maintaining AA
          instrumentation for graphite furnace,  flame,  and cold
          vapor AA,  or,  in lieu of the educational requirement,
          three additional years of experience in operating and
          maintaining AA instrumentation,  including graphite
          furnace,  flame,  and cold vapor  techniques.

4.   Inorganic Sample Preparation Specialist Qualifications

     a.    Education:

         Minimum of  high  school  diploma  and a college  level
         course in  general chemistry or  equivalent.

     b.   Experience:

         Minimum of  1 year of experience  in sample  preparation in
         an analytical  laboratory.

     c.   Experience  (Required if  microwave  digestion  is  used):

         Minimum of  six months experience  in an analytical
         laboratory  and six  months  experience  in sample
         dissolution using microwave  digestion  techniques.

5.   Classical Techniques  (Cyanide)  Analyst  Qualifications

     a.   Education:

         Minimum of  Bachelor's degree  in chemistry  or  any
         scientific/engineering discipline.

     b.   Experience:

         Minimum of  1 year of experience with classical  chemistry
         laboratory  procedures, in  conjunction  with the

                         A-10                                ILM01.3

-------
              educational qualifications; or,  in lieu of educational
              requirement, two years of additional equivalent
              experience.

     6.    Technical  Staff Redundancy

          In  order to  ensure  continuous operations to accomplish  the
          required work  as  specified by the  EPA contract, the bidder
          shall  have a minimum of one  (1)  chemist available  at  all
          times  as a back-up  technical person with the following
          qualifications.

          a.   Education:
              Minimum of Bachelor's degree  in  chemistry or any
              scientific/engineering discipline.

          b.   Experience:
              Minimum of one  year of experience in each of the
              following areas —

            ' o    ICP operation and maintenance
              o    AA operation and maintenance

              o    Classical chemistry analytical procedures

              o    Sample preparation for inorganics analysis
C.   Facilities

     The adequacy of the facilities and equipment is of  equal
     importance  for  the  technical staff to accomplish  the required  work
     as  specified by the EPA contract.

     1.    Sample  Receipt Area
          Adequate,  contamination-free, well-ventilated work space
          provided with  chemical  resistant bench top for receipt  and
          safe handling  of  EPA samples.

     2.    Storage Area
          Sufficient refrigerator space to maintain unused EPA  sample
          volume  for 60  days  after data submission.  Samples and
          standards  must be stored separately.
                              A-ll                                ILM01.0

-------
D.
3.   Sample Preparation Area
     Adequate,  contamination-free,  well-ventilated work space
     provided with:

     a.    Benches  with chemical  resistant  tops.

     b.    Exhaust  hoods.   Note:  Standards must be  prepared  in a
          glove box  or isolated  area.

     c.    Source of  distilled  or demineralized organic-free  water.
     d.    Analytical balance(s)  located  away  from  draft and  rapid
          change in  temperature.

Instrumentation

At a minimum, the Contractor shall have the following instruments
operative at the time of the Preaward Site Evaluation and
committed for the full duration of the contract.

1.   200 Samples/Month Capacity  Requirements
Fraction
ICP Metals
GFAA Metals
1
1
Mercury
Cyanide
No. of
Instrument(s)
1
2
2
12 distillation
units + 1
photometer
1
Type of |
Instrument |
1
ICP Emission
Spec tropho tome ter
Atomic Absorption
Spec tropho tome ter
with Graphite
Furnace Atomizer
Mercury Cold Vapor
AA Analyzer or AA
instrument
modified for Cold
Vapor Analysis
See Cyanide
Methods, Staterhent
of Work Exhibit D,
Section IV, Part E
There  are  no  Secondary Instrument Requirements for 200 Samples/Month
Capacity.
                               A-12
                                                              ILM01.0

-------
2.   300 Samples/Month Capacity Requirements
Fraction
ICP Metals
GFAA Metals
Mercury
Cyanide
No. of
Instrument(s)
1
3
2
12 distillation
units + 1
photometer
Type of
Instrument
ICP Emission
Spectrophotometer
Atomic Absorption
Spectrophotometer
with Graphite
Furnace Atomizer
Mercury Cold Vapor
AA Analyzer or AA
instrument
modified for Cold |
Vapor Analysis
See Cyanide
Methods, Statement
of Work Exhibit D,
Section IV, Part E
Secondary Instrument Requirements for 300 Samples/Month Capacity

     The Contractor shall have the following instruments in place
     and operational at any one time as a back-up system:
     Quantity
         One
Instruments
GFAA
3.   Additional Instrument Requirements for greater than 300
     Samples/Month Capacity
     Quantity

         One

         One
Instruments

GFAA

ICP Emission Spectrophotometer
4.   Instrument Specifications

     Further information on instrument specifications and required
     ancillary equipment may be found in the Statement of Work.
                         A-13
                          ILM01.0

-------
     E.   Data Management and Handling

          1.   Hardware   -  Contractor  will  have  an  IBM or  IBM-compatible
               mini-computer  or  PC  capable  of recording required sample data
               on 5.25 inch double-sided, double-density  360  K-byte or high
               density 1.2 M-byte  diskettes;  or a.  3.5  inch double-sided,
               double-density 720 K-byte or  1.44  M-byte  diskettes  in ASCII
               text  file  format  and in accordance with  the  file,  record and
               field specifications listed in  the  SOW,  Exhibit H.

               Other minimum  requirements include:

                   Hard  disk of at least 20 M-bytes.

                   Modem capable of  at least 2,400  baud transmission speed
                   which  is   compatible  with  the  EPA  Telecommunications
                   Network.

          ?     Software  -  Software,  utilized  in generating,  updating  and
               quality  controlling  analytical   databases    and   automated
               deliverables    shall    have    the    following    additional
               capabilities:

                   Editing and updating databases.

                   QC of automated deliverables.

                   Controlled  access  using  user   ID   and   file   password
                   protection.

          3.   The Contractor  shall also  be  able  to submit reports  and  data
              packages as specified in the SOW Exhibit  B.  To complete  this
              task,  the  Contractor  shall  be required  to  provide  space,
              tables  and  adequate  copy  machines   to  meet  the   contract
              requirements.

II.   LABORATORY MANAGEMENT CAPABILITY

     The   Contractor   must   have    an  organization   with   well-defined
     responsibilities  for each individual in the management system  to  ensure
     sufficient resources for EPA contract(s) and to maintain a successful
     operation.     To   establish  this  capability,   the  Contractor  shall
     designate  personnel  to carry out  the following responsibilities for the
     EPA   contract.    Functions  include,  but  are  not  limited   to,  the
     following:

     A.   Technical Staff

         Responsible  for all  technical efforts for the EPA contract.
                                  A-14                                ILM01.0

-------
B.    Project Manager

     Responsible  for overall aspects of EPA contract(s)  (from sample
     receipt through data delivery) and shall be the primary contact
     for EPA Headquarters Administrative Project Officer and Regional
     Technical Project Officers.

C.    Sample Custodian

     Responsible  for receiving the EPA samples (logging, handling and
     storage).

D.    Quality Assurance Officer

     Responsible  for overseeing the quality assurance aspects of the
     data and reporting directly to upper management.

E.    Data Reporting and Delivery Officer

     Responsible  for all aspects of data deliverables:   organization,
     packaging, copying, and delivery.
                              A-15                                ILM01.0

-------
                                  EXHIBIT B

                  REPORTING  AND DELIVERABLES REQUIREMENTS

                                                              Page No.

SECTION I:   Contract Reports/Deliverables Distribution ....   B-2

SECTION II:  Report Descriptions and Order of Data
             Deliverables  	   B-4

SECTION III: Form Instruction Guide 	   B-L5

SECTION IV:  Data Reporting  Forms  	   B-42
                                                                      ILM01.0

-------
                                        SECTION I

                        CONTRACT REPORTS/DELIVERABLES DISTRIBUTION

 The following table reiterates  the  Contract  reporting  and deliverables requirements
 specified in the Contract  Schedule  and  specifies the distribution that is required for
 each deliverable.   NOTE: Specific recipient  names and  addresses  are subject to change
 during the term of the  contract.  The Administrative Project  Officer will notify the
 Contractor in writing of such changes when they occur.
                             No.
Delivery
|    Distribution
Item

A. Updated SOPs

B. Sample Traffic
Reports



**C. Sample Data
Package

D. Data in Computer
Readable Format

****£. Complete SDG File


*F. Quarterly/ Annual
Verification
of Instrument
Parameters
*****G . Qua 1 i ty
Assurance
Plan

Copies

2

1




2


1


1


2
t


copy



Schedule

45 days after
contract receipt
3 days after
receipt of last
sample in Sample
Delivery Group
(SDG)***
35 days after
receipt of last
sample in SDG
35 days after
receipt of last
sample in SDG
35 days after
receipt of last
sample in SDG**
Quarterly:
15th day of
January, April
July, October
Submit copy
within 7 days
of written
request by APO
(D



X




X


X





X



Ai



(2)
1
X












X






» direct



(3)

X






X








X



;ed



Distribution:
(1)  Sample Management Office (SMO)
(2)  Region-Client
(3)  Environmental Monitoring Systems Laboratory (EMSL)
                                        B-2
                                    ILM01.0

-------
      *     Also required in each Sample Data Package.

      **    Concurrent delivery of these items to all recipients is required.

      ***   Sample Delivery Group (SDG) is a group of samples within a Case,
            received over a period of 14 days or less and not exceeding 20
            samples.   Data for all samples in the SDG are due concurrently.
            (See SOW Exhibit A, paragraph I., for further description).

      **•**  Complete SDG file will contain the original sample data package
            plus all of the original documents described in Exhibit B of the
            Statement of Work under Complete SDG File.

      *****See Exhibit E for description

     NOTE:   As specified in the Contract Schedule (Government Furnished
     Supplies  and Materials),  unless otherwise instructed by the CLP Sample
     Management Office,  the Contractor shall dispose of unused sample volume
     and used  sample  bottles/containers no earlier than sixty (60)  days
     following submission of analytical data.

Distribution Addresses:

(1)  USEPA Contract Laboratory Program (CLP)
     Sample Management Office (SMO)
     P.  0.  Box 818
     Alexandria, VA  22313
     For overnight delivery service, use street address:
     300 N. Lee Street
     Alexandria, VA  22313

(2)  USEPA REGIONS:  The CLP Sample  Management Office, acting on behalf of
     the Administrative  Project Officer, will provide the Contractor with the
     list of addressees  for the ten EPA Regions.   SMO will provide  the
     Contractor with updated Regional address/name lists as necessary
     throughout the period of the contract and identify other client
     recipients on a case-by-case basis.

(3)  USEPA Environmental Monitoring Systems Laboratory (EMSL)
     944 E. Harmon Avenue
     Las Vegas, NV  89109
                                    B-3                                 ILM01.0

-------
                                  SECTION II

              REPORT DESCRIPTIONS  AND ORDER OF DATA DELIVERABLES

      The Contractor laboratory shall provide reports and other deliverables
as specified in the Contract Performance/Delivery Schedule (see Contract
Schedule, Section F) .   The required content and form of each deliverable is
described in this Exhibit.

     All reports and documentation MUST BE

     o  Legible,

     o  Clearly labeled and completed  in  accordance with instructions in
        this Exhibit,

     o  Arranged in the order specified in  this Section,

     o  Paginated sequentially according  to  instructions in this Exhibit, and
     o  Single-sided.

      If submitted documentation does not conform to the above criteria, the
Contractor will be required to resubmit such documentation with
deficiency(ies) corrected, at  no additional cost to the government.

      The Contractor must be prepared to receive the full monthly sample
contract requirement at the time of contract award.

      Whenever the Contractor  is required to submit or resubmit data as a
result of an on-site laboratory evaluation or through an APO/TPO action, the
data must be clearly marked as ADDITIONAL DATA and must be  sent to all three
contractual data recipients (SMO,  EMSL, and Region).   A cover  letter shall be
included which describes what  data is being delivered,  to which EPA Case(s)
the data pertains,  and who requested the data.

      Whenever the Contractor  is required to submit or resubmit data as a
result of Contract Compliance  Screening (CCS) review by SMO,  the data must be
sent to all three contractual  data recipients (SMO, EMSL and Region),  and in
.11 three instances must be accompanied by a color-coded COVER SHEET
(Laboratory Response To Results of Contract Compliance Screening) provided by
SMO.  Diskette deliverables need only be submitted or resubmitted to SMO.

      Section IV of this Exhibit contains the required Inorganic Analysis
Data Reporting Forms in Agency-specified formats;  Section III  of this Exhibit
contains instructions to the Contractor for properly completing all data
reporting forms to provide the Agency with all required data.   Data elements
and field descriptors for reporting data in computer-readable  format are
contained in Exhibit H.

      Descriptions of the requirements for each deliverable item cited in the
Contract Performance/Delivery  Schedule (see Contract Schedule,  Section F)  are
specified in parts A-G of this Section.  Items submitted concurrently must be
arranged in the order listed.   Additionally, the components of each item must
be arranged in the order presented herein when the item is  submitted.
                                   B-4                                 ILM01.0

-------
A.   Updated SOPs

     The Contractor shall submit updated copies of all required Standard
     Operating Procedures (SOPs) that were submitted with the prebid
     Performance Evaluation sample results.   The updated SOPs must address
     any and all issues of laboratory performance and operation identified
     through the review of the Performance Evaluation sample data and the
     evaluation of Bidder-Supplied Documentation.

     The Contractor must supply SOPs for the following:

          1.    Sample receipt  and  logging.

          2.    Sample storage.

          3.    Preventing  sample contamination.

          4.    Security  for laboratory and samples.

          5.    Standards purity/preparation.

          6.   Maintaining instrument records and logbooks.

          7.    Sample analysis and data control systems.

          8.   Glassware cleaning.

          9.   Technical and managerial review of laboratory operation and
               data package preparation.

          10.   Internal  review of contractually-required quality assurance and
              quality control data for each individual data package.

          11.   Sample analysis, data handling and reporting.

          12.  Chain-of-custody procedures and document control including SDG
              file preparation

          13.  Sample data validation/Self-inspection system

              a.   Data flow and chain-of-command for data review

              b.   Procedures for measuring precision and accuracy

              c.   Evaluation parameters for identifying systematic errors

              d.   Procedures to assure that hardcopy and diskette
               j    deliverables are complete and compliant with the
                   requirements in Exhibits  B and H

              e.   Procedures to assure that hardcopy deliverables are  in
                   agreement with their comparable diskette deliverables
                                   B-5                                ILM01.0

-------
               f.   Demonstration of internal QA inspection procedure
                   (demonstrated by supervisory sign-off on personal
                   notebooks, etc.)

               g.   Frequency and type of internal audits (e.g., random,
                   quarterly, spot checks, perceived trouble areas)

               h.   Demonstration of problem identification-corrective actions
                   and resumption of analytical processing.  Sequence
                   resulting from internal audit (i.e., QA feedback)

               i.   Documentation of audit reports (internal and external),
                   response, corrective action, etc.

          14.   Data Management and Handling

               a.   Procedures for controlling and estimating data entry
                   errors.

               b.   Procedures for reviewing changes to data and deliverables
                   and enduring traceability of updates.

               c.   Lifecycle management procedures for testing, modifying and
                   implementing changes to existing computing systems
                   including hardware, software, and documentation or
                   installing new systems.

               d.   Database security, backup and archival procedures
                   including recovery from system failures.

               e.   System maintenance procedures and response time.

               f.   Individual(s) responsible for system operation,
                   maintenance, data integrity and security.

               g.   Specifications for staff training procedures.

B.   Sample Traffic  Reports

      Original Sample Traffic Report page marked "Lab Copy for Return to SMO"
      with lab receipt information and signed in original Contractor
      signature, shall be submitted for each sample in the Sample Delivery
      Group.

      Traffic Reports (TRs) shall be submitted in Sample Delivery Group (SDG)
      sets (i.e., TRs for all samples in an SDG shall be clipped together),
      with an  SDG Cover Sheet attached.

      The SDG  Cover Sheet shall contain the following items:

      o  Lab  name
      o  Contract number
                                    B-6                                 ILM01.0

-------
      o Sample Analysis  Price  -  full  sample price from contract.
      o Case Number
      o List of EPA sample  numbers  of all  samples in the SDG, identifying  the
        first and  last samples received, and  their dates of receipt.
      NOTE: When more than one sample is received in the first or last SDG
      shipment,  the "first" sample received would be the lowest sample number
      (considering both alpha and numeric designations);  the "last" sample
      received would be the highest sample number (considering both alpha and
      numeric designations).

      In addition,  each Traffic Report must be clearly marked with the SDG
      Number, the sample number of the first sample in the SDG (as described
      in the following paragraph).   This information should be entered below
      the Lab Receipt Date on the TR.

      EPA field sample numbers are six digits in length.   If the Contractor
      receives sample numbers of any other length,  contact SMO immediately.
      The EPA sample number of the first sample received in the SDG is the
      SDG number.   When several samples are received together in the first
      SDG shipment, the SDG number shall be the lowest sample number
      (considering both alpha and numeric designations)  in the first group of
      samples received under the SDG.   (The SDG number is also reported on
      all data reporting forms.  See Section III, Form Instruction Guide.)

      If samples are received at the laboratory with multi-sample Traffic
      Reports (TRs),  all the samples on one multi-sample TR may not
      necessarily be in the same SDG.   In this instance,  the laboratory must
      make the appropriate number of photocopies of the TR, and submit one
      copy with each SDG cover sheet.

C.    Sample Data Package

      The sample data package shall include data for analysis of all samples
      in one Sample Delivery Group (SDG), including field and analytical
      samples,  reanalyses, blanks,  spikes,  duplicates,  and laboratory control
      samples.

      The sample data package must be complete before submission,  must be
      consecutively paginated (starting with page number one and ending with
      the number of all pages in the package), and shall include the
      following:

     1.    Cover  Page  for  the Inorganic Analyses  Data  Package,  (COVER  PAGE  --
          Inorganic Analyses Data Package),  including:   laboratory  name;
          laboratory  code;  contract  number;  Case No.; Sample  Delivery Group
          (SDG)  No.;  SAS  Number (if  appropriate); EPA sample  numbers  in
          alphanumeric  order,  showing  EPA sample numbers  cross-referenced with
          lab ID numbers;  comments,  describing in detail  any  problems
          encountered in  processing  the samples  in  the data package;  and,
          completion  of the  statement  on use  of  ICP background and
          interelement  corrections  for the  samples.
                                    B-7                                 ILM01.0

-------
     The Cover Page  shall  contain the  following statement,  verbatim:   "I
     certify that this  data package  is in compliance  with the terms and
     conditions of the  contract,  both  technically and for completeness,
     for other than  the conditions detailed above.  Release of the data
     contained in this  hardcopy data package and in the computer-readable
     data submitted  on  diskette has  been authorized by the Laboratory
     Manager or the  Manager's  designee,  as verified by the following
     signature."  This  statement shall be directly followed by the
     signature of the Laboratory Manager or his designee with a typed
     line below it containing  the signers name  and title,  and the date of
     signature.

     In addition, on a  separate piece  of paper,  the Contractor must also
     include any problems  encountered;  both technical and administrative,
     the corrective  action taken and resolution.

2.   Sample  Data

     Sample  data shall  be  submitted  with the Inorganic Analysis Data
     Reporting Forms for all samples in the SDG,  arranged in increasing
     alphanumeric EPA sample number  order,  followed by the  QC analyses
     data, Quarterly Verification of Instrument Parameters  forms,  raw
     data, and copies of the digestion and distillation logs._

     a.   Results -- Inorganic Analysis  Data Sheet  [FORM I  -  IN]

         Tabulated  analytical results  (identification and  quantitation)
         of the specified  analytes  (Exhibit C).  The  validation and
         release of these  results is  authorized by a  specific,  signed
         statement  on  the  Cover  Page.   If  the Laboratory Manager cannot
         verify all data reported for each sample, he/she  must provide  a
         detailed description of the  problems associated with  the
         sample(s)  on  the  Cover  Page.

         Appropriate concentration  units  must be specified and entered
         on Form I.  The quantitative values shall be reported in units
         of micrograms  per liter (ug/L)  for aqueous  samples  and
         milligrams per kilogram (mg/kg)  for solid samples.  No other
         units  are  acceptable. Results  for solid samples must  be
         reported on a  dry weight basis.   Analytical  results must be
         reported to two significant  figures if  the  result value  is  less
         than 10; to three significant  figures  if  the value  is greater
         than or equal  to  10.  Results  for percent solids  must be
         reported to one decimal place.   The preceding discussion
         concerning significant  numbers  applies  to Form I  only.   For
         other Forms,  follow the instructions specific  to  those forms as
         contained  in  this exhibit.

     b.   Quality Control Data

          1)   Initial and  Continuing Calibration Verification   [FORM II
               (PART 1)  - IN]

          2)   CRDL  Standard for  AA  and  Linear Range Analysis for  ICP
               [FORM II  (PART 2)  - IN]


                               B-8                                 ILM01.0

-------
      3)   Blanks [FORM III - IN]

      4)   ICP Interference Check Sample [FORM IV - IN]

      5)   Spike Sample Recovery [FORM V (PART 1) - IN]

      6)   Post Digest Spike Sample Recovery [FORM V (PART 2) - IN]

      7)   Duplicates [FORM VI - IN]

      8)   Laboratory Control Sample  [FORM VII - IN]

      9)   Standard Addition Results  [FORM VIII - IN]

     10)   ICP Serial Dilutions [FORM IX -  IN]

     11)   Preparation Log [Form XIII -  IN]

     12)   Analysis Run Log [Form XIV -  IN]

c.   Quarterly Verification of Instrument Parameters

      1)   Instrument Detection Limits (Quarterly)  [FORM X -  IN]

      2)   ICP Interelement Correction Factors  (Annually)  [FORM XI
          (PART 1)  - IN]

      3)   ICP Interelement Correction Factors  (Annually)  [FORM XI
          (PART 2)  - IN]

      4)   ICP Linear Ranges (Quarterly)  [FORM  XII  - IN]

     (Note that copies of Quarterly  Verification of Instrument
     Parameters forms for the current quarter  must be  submitted with
     each data package.)

d.   Raw  Data

     For  each reported value, the  Contractor  shall include in the
     data package all raw data  used to obtain that value.   This
     applies to all required QA/QC measurements,  instrument
     standardization, as  well as all sample analysis results.   This
     statement does not apply to the Quarterly Verification  of
     Instrument Parameters submitted as a  part of  each data  package.
     Raw  data must contain all instrument  readouts used for  the
     sample results.  Each exposure  or  instrumental reading  must  be
     provided,  including  those readouts that  may fall  below  the  IDL.
     All  AA and ICP instruments must provide  a legible hard  copy  of
     the  direct real-time instrument readout  (i.e., stripcharts,
     printer tapes,  etc.).   A photocopy of the instruments direct
     sequential readout must be included.   A hardcopy  of the
     instrument's direct  instrument  readout for cyanide must be
     included if the instrumentation has the  capability.
                          B-9                                 ILM01.0

-------
The order of raw data in the data package shall be:  ICP, Flame
AA, Furnace AA, Mercury, and Cyanide.  All raw data shall
include concentration units for ICP and absorbances or
concentration units for flame AA, furnace AA,  Mercury and
Cyanide.  All flame and furnace AA data shall be grouped by
element.

Raw data must be labeled with EPA sample number and appropriate
codes, shown in Table 1 following, to unequivocally identify:

 1)   Calibration standards,  including source and prep date.

 2)   Initial and continuing calibration blanks and preparation
     blanks.

 3)   Initial and continuing calibration verification standards,
     interference check samples,  ICP serial  dilution samples,
     CRDL Standard for ICP and AA, Laboratory  Control Sample
     and Post Digestion Spike.

 4)   Diluted and undiluted samples (by EPA sample number) and
     all weights, dilutions  and volumes used to obtain the
     reported values.   (If the volumes, weights and dilutions
     are consistent for all  samples in a given SDG, a general
     statement outlining these parameters is sufficient).

 5)   Duplicates.

 6)   Spikes (indicating standard  solutions used,  final spike
     concentrations,  volumes involved).  If  spike information
     (source, concentration,  volume)  is consistent for a  given
     SDG, a general statement outlining these  parameters  is
     sufficient.

 7)   Instrument used,  any instrument adjustments,  data
     corrections or other apparent anomalies on the measurement
     record, including all data voided or data not used to
     obtain reported values  and a brief written explanation.

 8)   All information for furnace  analysis clearly and
     sequentially identified on the raw data,  including EPA
     sample number,  sample and analytical spike data,  percent
     recovery,  coefficient of variation,  full  MSA data, MSA
     correlation coefficient,  slope and intercepts of linear
     fit, final sample concentration (standard addition
     concentration),  and type of  background  correction used:
     BS for Smith-Heiftje,  BD for Deuterium  Arc,  or BZ for
 ,    Zeeman.

 9)   Time and date of each analysis.   Instrument run logs can
     be submitted if they contain this information.  If the
     instrument does not automatically provide times of
     analysis,  these must be manually entered  on all raw  data
                     B-10                               ILM01.0

-------
                   for initial and continuing calibration verification and
                   blanks, as well as  interference check samples and CRDL
                   standard for  ICP.

              10)  Integration times for AA analyses.

          e.   Digestion  and Distillation Logs

              Logs shall be submitted  in the following  order:  digestion logs
              for ICP, flame AA,  furnace AA and mercury preparations,
              followed by a copy of  the distillation log for  cyanide.  These
              logs must  include:   (1)  date, (2) sample  weights and volumes,
              (3) sufficient information to unequivocally  identify which QC
              samples (i.e., laboratory control sample, preparation blank)
              correspond to each batch digested,  (4)  comments describing any
              significant sample changes or reactions which occur during
              preparation, and  (5) indication of  pH  <2  or  >12,  as applicable.

     3.    A copy  of the Sample Traffic  Reports submitted in Item A for  all  of
          the  samples  in  the SDG.  The  Traffic Reports shall be arranged in
          increasing  EPA  Sample  Number  order, considering both alpha and
          numeric designations.   A legible  photocopy  of the SDG cover sheet
          must also be submitted.

D.   Data in Computer Readable  Form

      The Contractor shall provide a computer-readable copy of the data on
      data reporting Forms I-XIV  for all samples  in the Sample Delivery
      Group, as specified in the  Contract Performance/Delivery Schedule.
      Computer-readable data deliverables shall be submitted on an IBM or
      IBM-compatible, 5.25 inch  floppy double-sided,  double density 360 K-
      byte or a high density 1.2  M-byte diskette  or on an IBM or IBM-
      compatible, 3.5 inch double-sided,  double density 720 K-byte or a high
      density 1.44 M-byte diskette.  The data shall be recorded in ASCII,
      text file format,  and shall adhere to the file, record and field
      specifications listed in Exhibit H,  Data Dictionary and Format for Data
      Deliverables in Computer-Readable Format.

      When submitted, diskettes  shall be packaged and shipped in such a
      manner  that the diskette(s) cannot be bent  or folded, and will not be
      exposed to extreme heat or  cold or any type of electromagnetic
      radiation.   The diskette(s) must be included in the same shipment as
      the hardcopy data and shall, at a minimum,  be enclosed in a diskette
      mailer.
                                    B-ll                                ILM01.0

-------
                                   Table 1

                           Codes for Labelling Data
      Sample                                                   XXXXXX
      Sample not part of the SDG                               ZZZZZZ
      Duplicate                                               XXXXXXD
      Matrix Spike                                            XXXXXXS
      Serial Dilution                                         XXXXXXL
      Analytical Spike                                        XXXXXXA
      Post Digestion/Distillation Spike                       XXXXXXA
      MSA:
         Zero Addition                                         XXXXXXO
         First Addition                                        XXXXXX1
         Second Addition                                       XXXXXX2
         Third Addition                                        XXXXXX3
      Instrument Calibration Standards:
         ICP                                S  or  SO  for blank standard
         Atomic Absorption  and Cyanide                 SO, S10,...etc.
         Initial Calibration Verification                          ICV
         Initial Calibration Blank                                 ICB
      Continuing Calibration Verification                         CCV
      Continuing Calibration Blank                                CCB
      Interference Check Samples:
         Solution A                                               ICSA
         Solution AB                                              ICSAB
      CRDL Standard for AA                                        CRA
      CRDL Standard for ICP                                       CRI
      Laboratory Control Samples:
         Aqueous (Water)                                           LCSW
         Solid  (Soil/Sediment)                                     LCSS
      Preparation Blank (Water)                                    PBW
      Preparation Blank (Soil)                                    PBS
      Linear Range Analysis Standard                              LRS
Notes:

1.   When an analytical spike or MSA is performed on  samples other  than  field
     samples, the "A",  "0",  "1", "2" or "3"  suffixes  must be the  last  to be
     added to the EPA Sample Number.  For instance, an  analytical spike  of  a
     duplicate must be  formatted "XXXXXXDA."

2.   The numeric suffix that follows the "S"  suffix for the standards
     indicates the true value of the concentration of the standard  in  ug/L.
                                   B-12                                ILM01.0

-------
3.   ICP calibration standards usually consist of several analytes at
     different concentrations.  Therefore,  no numeric suffix can follow the
     ICP calibration standards unless all the analytes in the standard are
     prepared at the same concentrations.  For instance,  the blank for ICP
     must be formatted "SO."
4.   Use suffixes of "0", "1", "2",  "3" as  appropriate for samples identified
     with ZZZZZZ on which MSA has been performed to indicate single
     injections.
E.   Results of Intercomparison/Performance  Evaluation (PE)  Sample Analyses

      Tabulation of analytical results for Intercomparison/PE Sample analyses
      include all requirements specified in items C.  and D.,  above.

F.   Complete SPG File (CSF)

     As  specified in the  Delivery  Schedule,  one  Complete SDG  File (CSF)
     including the original Sample Data Package  shall  be delivered to the
     Region concurrently  with delivery of a  copy of the  Sample  Data Package
     to  SMO and EMSL/LV.   The contents of the CSF will be numbered according
     to  the specifications described  in Section  III and  IV of Exhibit B.   The
     Document Inventory Sheet,  Form DC-2,  is contained in Section IV.  The
     CSF will contain all original documents where possible.  No  copies  of
     original documents will  be placed in the CSF unless the  originals are
     bound in a logbook maintained by the laboratory.  The CSF  will contain
     all original documents specified in Section III and IV,  and  Form DC-2 of
     Exhibit B of the SOW.

     The CSF will consist of  the following original documents in  addition to
     the documents in the Sample Data Package:

     1.    Original  Sample  Data  Package

     2.    A completed and  signed Document  Inventory Sheet (Form DC-2)

     3.    All  original  shipping  documents, including,  but not limited  to,  the
          following  documents:

          a.   EPA Chain-of-Custody Record

          b.   Airbills

          c.   EPA (SMO) Traffic Reports

          d.   Sample Tags (if present) sealed  in plastic bags.

     4.    All  original receiving documents, including,  but not limited to,  the
          following documents:

          a.   Form DC-1

          b.   Other receiving forms or copies  of receiving  logbooks.
                                  B-13                                ILM01.0

-------
         c.   SDG Cover Sheet

     5.   All original laboratory records of sample transfer, preparation, and
         analysis, including, but not limited to, the following documents:

         a.   Original preparation and analysis forms or copies of
              preparation and analysis logbook pages.

         b.   Internal sample and sample digestate/distillate transfer chain-
              of-custody records.

     6.   All other original case-specific documents in the possession of  the
         laboratory, including, but not limited to, the following documents:

         a.   Telephone contact logs.

         b.   Copies of personal logbook pages.

         c.   All handwritten case-specific notes.

         a.   Any other case specific documents not covered by the above.

         NOTE:  All case-related documentation may be used or admitted as
         evidence in subsequent legal proceedings.  Any other case-specific
         documents generated after the CSF is sent to EPA, as well as copies
         that are altered in any fashion, are also deliverables to EPA
         (original to the Region and copies to SMO and EMSL/LV).

         If the laboratory does submit case-specific documents to EPA after
         submission of the CSF, the documents should be numbered as an
         addendum to the CSF and a revised DC-2 form should be submitted; or
         the documents should be numbered as a new CSF and a new DC-2 form
         should be submitted to the Regions only.

G.   Quarterly  and Annual Verification of Instrument Parameters

     The  Contractor shall perform and report quarterly verification of
     instrument detection limits and  linear ran^e  methods  specified in
     Exhibit E  for each  instrument used under this contract.  For  the ICP
     instrumentation,  the Contractor  shall also perform and  report annual
     interelement correction  factors  (including method of  determination),
     wavelengths used  and integration times.  Forms for Quarterly  and Annual
     Verification of Instrument  Parameters for the current quarter and year
     shall  be  submitted  in  each  SDG data package,  using Forms X, XIA, XIB,
     and XII.   Submission of  Quarterly/Annual Verification of Instrument
     Parameters  shall  include the raw data used to determine those values
     reported.
                                   B-14                                ILM01.0

-------
                                  SECTION  III


                            FORM INSTRUCTION GUIDE


This section contains specific instructions for the completion of all
required Inorganic Data Reporting Forms.  This section is organized into the
following Parts:

     A.    General  Information and Header Information

     B.    Cover Page  --  Inorganic Analyses Data Package [COVER PAGE -  IN]

     C.    Inorganic Analysis Data Sheet  [FORM I  -  IN]

     D.    Initial  and Continuing  Calibration Verification  [FORM  II  (PART  1)  -
          IN]

     E.    CRDL Standard  Jor AA and ICP  [FORM  II  (PART  2,  -  IN]

     F.    Blanks  [FORM  III -  IN]

     G.    ICP  Interference Check  Sample  [FORM IV  -  IN]

     H.    Spike Sample Recovery  [FORM V (PART 1)  -  IN]

     I.    Post Digest Spike Sample Recovery  [FORM  V (PART 2)  - IN]

     J.    Duplicates   [FORM VI -  IN]

     K.    Laboratory  Control Sample  [FORM VII -  IN]

     L.    Standard Addition Results  [FORM VIII - IN]

     M.    ICP  Serial  Dilutions  [FORM IX -  IN]

     N.    Instrument  Detection Limits (Quarterly)   [FORM  X - IN]

     0.    ICP  Interelement Correction Factors  (Annually)   [FORM XI
          (PART 1)  -  IN]

     P.    ICP  Interelement Correction Factors  (Annually)   [FORM XI
          (PART 2)  -  IN]

     Q.    ICP  Linear  Ranges (Quarterly)   [FORM XII  -  IN]

     R.    Preparation Log   [Form  XIII -  IN]

     S.    Analysis Run Log  [Form XIV -  IN]

     T.    Sample Log-In  Sheet   [Form  DC-1]

     U.    Document Inventory Sheet  [Form  DC-2]


                                   B-15                                ILM01.0

-------
A.   General Information and Header Information

     The data reporting forms presented of Section IV in this Exhibit have
     been designed in conjunction with the computer-readable data format
     specified in Exhibit H,  Data Dictionary and Format for Data Deliverables
     in Computer-Readable Format.  The specific length of each variable for
     computer-readable data transmission purposes is given in the Data
     Dictionary (Exhibit H).   Information entered on these forms must not
     exceed the size of the field given on the  form, including such
     laboratory-generated items  as Lab Name and Lab Sample ID.

     Note that on the hardcopy forms (see Section IV),  the space provided for
     entries is greater in some  instances than  the length prescribed for the
     variable as written to diskette (see Exhibit H).  Greater space is
     provided on the hardcopy forms for the sake of visual clarity.

     Values must be reported on  the hardcopy forms according to the
     individual form instructions in this Section.   Each form submitted muse
     be filled out completely for all analytes  before proceeding to the next
     form of the same type.   Multiple forms cannot be submitted in place of
     one form if the information on those forms can be submitted on one form.

     All characters which appear on the data reporting forms presented in the
     contract (Exhibit B,  Section IV)  must be reproduced by the Contractor
     when submitting data,  and the format of the forms submitted must be
     identical to that shown in  the contract.   No information may be added,
     deleted, or moved from its  specified position without prior written
     approval of the EPA Administrative Project Officer.   The names of the
     various fields and analytes (i.e.,  "Lab Code",  "Aluminum") must appear
     as they do on the forms  in  the contract, including the options specified
     in the form (i.e.,  "Matrix  (soil/water):"  must appear,  not just
     "Matrix").

     All alphabetic entries made onto the forms by the  Contractor must be in
     ALL UPPERCASE letters  (i.e.,  "LOU",  not "Low"  or "low").   If an entry
     does not fill the entire blank space provided on the  form,  null
     characters must be used to  remove the remaining underscores that
     comprise the blank line.  (See Exhibit H for more  detailed
     instructions.)  However, do not remove the underscores or  vertical bar
     characters that delineate "boxes" on the forms.

     Six pieces of information are common to the header sections of each data
     reporting form.   These are:   Lab Name,  Contract,  Lab  Code,  Case No.,  SAS
     No.,  and SDG No.   This information must be entered on every form and
     must match on all forms.

     The "Lab Name" must be the  name chosen by  the  Contractor to identify the
     laboratory.  It may not exceed 25 characters.

     The "Contract" is the  number of the EPA contract under which the
     analyses were performed.

     The "Lab Code" is an alphabetic abbreviation of up to 6 characters,
     assigned by EPA, to identify the laboratory and aid in data processing.

                                   B-16                                ILM01.0

-------
 This lab code shall be assigned by EPA at  Che  time a contract is
 awarded, and must not be modified by the Contractor,  except at the
 direction of EPA.

 The "Case No."  is the EPA-assigned Case number (to 5 spaces)  associated
 with the sample,  and reported on the Traffic Report.

 The "SAS No." is  the EPA-assigned number for analyses performed under
 Special  Analytical Services.   If samples are to be analyzed under  SAS
 only,  and reported on these  forms,  then enter  SAS  No.  and leave Case No.
 blank.   If samples are analyzed according  to this  SOW (Routine
 Analytical Services protocol)  and have additional  SAS requirements,  list
 both Case No. and SAS No.  on all forms.  If the analyses  have no SAS
 requirements, leave "SAS No."  blank.   (NOTE:   Some samples in an SDG may
 have a SAS No., while others  do not.)

 The "SDG No." is  the Sample  Delivery Group (SDG) number.   The SDG  number
 is  the EPA Sample  Number of  the first sample received in  the  SDG.  When
 several  samples are received together in the first SDG shipment,the SDG
 number must be the lowest sample number (considering  both alpha and
 numeric  designations)  in the  first  group of samples  received  under the
 SDG.

 The other information common to several of the  forms  is the  "EPA Sample
 No.".  This  number appears either in the upper  righthand  corner of the
 form, or as  the left column  of a table summarizing data from  a number of
 samples.   When "EPA Sample No."  is  entered into the  triple-spaced  box in
 the upper righthand corner of  a form,  it must be centered on  the middle
 line of  the  three  lines  that comprise  the  box.

 All samples,  matrix spikes and duplicates  must  be  identified  with  an EPA
 Sample Number.  For  samples, matrix  spikes and  duplicates,  the EPA
 Sample Number is the  unique identifying number  given  in the Traffic
 Report that  accompanied  that sample.

 In  prder to  facilitate data assessment, the sample suffixes listed in
 Table 1  must be used.

 Other pieces  of information are common to many  of  the Data Reporting
 Forms.   These include:  Matrix and Level.

 For  "Matrix", 'enter  "SOIL" for soil/sediment  samples, and enter  "WATER"
 for water  samples.  NOTE:  The matrix must be spelled out.
Abbreviations such  as  "S" or "W" must not be  used.

For  "Level", enter  the determination of concentration level.   Enter as
 "LOW" or  "MED",  not  "L" or "M".

Note:  All results must be transcribed to Forms II-XIV from the raw data
to  the specified number of decimal places  that  are  described  in Exhibit
B and Exhibit H.   The raw data result is to be  rounded only when the
number of  figures in the raw data result exceeds the maximum number of
figures  specified for that result entry for that form.  If there are  not
enough figures in the raw data result to enter  in the specified space

                               B-17                                ILM01.0

-------
     for  that  result,  then zeros  must be used  for  decimal  places  to  the
     specified number  of reporting  decimals  for  that  result  for a specific
     form.   The following examples  are provided:

     Raw  Data  Result	Specified Format	Correct Entry on Form
95.99653
95.99653
95.99653
95.996
95.9
5.4 (to four decimal places)
5.3 (to three decimal places)
5.2 (to two decimal places)
5.4 (to four decimal places)
5.4 (to four decimal places)
95.9965
95.997
96.00
95.9960
95.9000
     For rounding off  numbers  to  the appropriate  level  of  precision,  observe
     the following common rules.   If the figure following  those  to be
     retained is  less  than 5,  drop it  (round down).   If the  figure is greater
     than 5,  drop it and increase the  last  digit  to be  retained  by 1  (round .
     up).   If the figure following the last digit to  be retained equals 5 and
     there are no digits to the right  of the 5 or all digits  to  the right of
     the 5 equals zero,  then round up  if the digit to be retained is  odd,  or
     round down if that  digit  is  even.  See also  Rounding  Rules  entry in
     Glossary (Exhibit G).

     Before evaluating a number for being in control  or out  of control of a
     certain limit, the  number evaluated must be  rounded using EPA rounding
     rules to the significance reported for that  limit.  For  instance, the
     control limit for an ICV  is  plus  or minus 10% of the  true value.   A
     reported percent  recovery value of 110.4 would be  considered in control
     while a reported  value of 110.6 would  be considered out  of  control.   In
     addition, a calculated value of 110.50 would be  in control  while a
     calculated value  of 110.51 would  be out of control.

B.   Cover Paee - Inorganic Analyses Data Package [COVER PAGE-IN]

     This form is used to list all samples  analyzed within a  Sample Delivery
     Group, and to provide certain analytical information  and general
     comments.  It is  also the document which is  signed by the Laboratory
     Manager to authorize and release  all data and deliverables  associated
     with the SDG.

     Complete the header information according to the instructions in Part A.

     For samples  analyzed using this SOW, enter "3/90"  for SOW No.

     Enter the EPA Sample No.  (including spikes and duplicates)  (to seven
     spaces) of every  sample analyzed  within  the  SDG.  Spikes must contain an
     "S" suffix and duplicates a  "D" suffix.  These sample numbers must be
     listed on the form  in ascending alphanumeric order.  Thus,  if MAB123 is
     the lowest (considering both alpha and numeric characters)  EPA Sample
     No. within the SDG, it would be entered  in the first  EPA Sample No.
     field.  Samples would be listed below  it, in ascending  sequence -
     MAB124, MAB125, MAC111, MAllll, MA1111D, etc.
                                   B-18                                ILM01.0

-------
     A maximum of twenty (20)  sample numbers can be entered on this form.
     Submit additional Cover Pages,  as appropriate, if the total number of
     samples,  duplicates,  and spikes in the SDG is greater than twenty (20).

     A Lab Sample ID (to ten spaces) may be entered for each EPA Sample No.
     If a Lab  Sample ID is entered,  it must be entered identically (for each
     EPA Sample No.) on all associated data.

     Enter "YES" or "NO" in answer to each of the two questions concerning
     ICP corrections.   Each question must be explicitly answered with a "YES"
     or a "NO".  The third question must be answered with a "YES" or "NO" if
     the answer to the second question is "YES".  It should be left blank if
     the answer to the second question is "NO".

     Under "Comments", enter any statements relevant to the analyses
     performed under the SDG as a whole.

     Each Cover Page must be signed, in original,  by the Laboratory Manager
     or the Manager's designee and dated, to authorize the release and verify
     the contents of all data and deliverables associateu with an SDG.

C.    Inorganic Analysis Data Sheet [FORM 1-IN]

     This form is used to tabulate and report sample analysis results for
     target analytes (Exhibit C).

     Complete  the header information according to the instructions in Part A
     and as follows.

     "Date Received" is the date (formatted MM/DD/YY) of sample receipt at
     the laboratory, as recorded on the Traffic Report, i.e., the Validated
     Time of Sample Receipt (VTSR).

     "% Solids" is the percent of solids on a weight/weight basis in the
     sample as determined by drying the sample as specified in Exhibit D.
     Report percent solids to one decimal place (i.e., 5.3%).  If the percent
     solids is not required because the sample is fully aqueous or less than
     1% solids, then enter "0.0".

     Enter the appropriate concentration units (UG/L for water or MG/KG dry
     weight for soil) .  Entering "MG/KG"  means "mg/Kg dry weight" on this
     form.

     Under the column labeled "Concentration", enter for each analyte either
     the value of the result (if the concentration is greater than or equal
     to the Instrument Detection Limit) or the Instrument Detection Limit for
     the analyte corrected for any dilutions (if the concentration is less
     than the  Instrument Detection Limit).

     Under the columns labeled "C",  "Q",  and "M",  enter result qualifiers as
     identified below.  If additional qualifiers are used, their explicit
     definitions must be included on the Cover Page in the Comments section.
                                   B-19                                ILM01.0

-------
 FORM  I-IN  includes  fields  for  three  types of  result  qualifiers.   These
 qualifiers must be  completed as  follows:

 o   C (Concentration) qualifier -- Enter "B" if the reported value was
    obtained from a reading that was less than the Contract Required
    Detection Limit (CRDL) but greater than or equal to the Instrument
    Detection Limit (IDL).  If the analyte was analyzed for but not
    detected, a "U" must be entered.

 o   Q qualifier -- Specified entries and their meanings are as follows:
    E   -     The reported value is estimated because  of  the presence of
             interference.  An explanatory note must be  included under
             Comments on the Cover Page (if the problem  applies to  all
             samples) or on the specific FORM I-IN  (if it is an isolated
             problem).

    M   -     Duplicate injection precision not met.

    N   -     Spiked  sample recovery not within control limits.

    S    -    The reported value was determined by the Method of Standard
            Additions (MSA).

    W   -     Post-digestion spike for Furnace AA analysis is out of
             control limits (85-115%), while sample absorbance is less
             than 50% of spike absorbance.  (See Exhibit E.)
    *   -    Duplicate analysis not within control limits.

    +   -     Correlation coefficient for the MSA is less than 0.995.

     Entering "S",  "W" ,  or "+"  is mutually exclusive.  No combination of
     these qualifiers can appear  in the same field for an analyte.

o   M (Method)  qualifier --  Enter:

         "P"  for ICP
         "A"  for Flame AA
         "F"  for Furnace AA
         "PM" for ICP when Microwave Digestion is used
         "AM" for flame AA when Microwave Digestion is used
         "FM" for Furnace AA when Microwave Digestion  is used
         "CV" for Manual Cold Vapor AA
         "AV" for Automated Cold Vapor AA
         "CA" for Midi-Distillation spectrophotometric.
         "AS" for Semi-Automated Spectrophotometric
         "C"  for Manual Spectrophotometric
         "T"  for Titrimetric
         "  "  where no data has been entered.
         "N"R" if the analyte is not required to be analyzed.

A brief physical description of the sample,  both before and after
digestion, must be  reported in the fields for color (before and after),
clarity (before and after), texture and artifacts.   For water samples,
report color and clarity.   For soil samples,  report color,  texture and
artifacts.
                              B-20                                ILM01.0

-------
 The  following  descriptive terms are recommended:

       Color         -   red, blue, yellow,  green,  orange, violet, white,
                        colorless, brown,  grey,  black
       Clarity       -   clear,  cloudy,  opaque
       Texture       -   fine  (powdery),  medium  (sand), coarse  (large
                        crystals or  rocks)

 If artifacts are  present, enter "YES"  in the  artifacts  field  and
 describe  the artifacts  in the  Comments field.   If artifacts are not
 present,  leave  this  field blank.

 Note any  significant changes  that occur during sample preparation  (i.e.,
 emulsion  formation)  in  the Comments field.  Enter any sample-specific
 comments  concerning  the analyte results in the Comments  field.

 Initial and Continuing  Calibration  Verification [FORM II(PART 1)-IN]

 This form is used to report  analyte recoveries from calibration
 solutions.

 Complete  the header  information according to  the  instructions  in Part A
 and as follows.

 Enter  the Initial Calibration  Source (12  spaces maximum) and  the
 Continuing Calibration  Source  (12 spaces  maximum).  Enter "EPA-LV" or
 "EPA-CI"  to indicate EPA EMSL  Las Vegas or Cincinnati, respectively, as
 the source of EPA standards.   When  additional  EPA supplied solutions are
 prepared  in the future,  the Contractor must use the codes supplied with
 those  solutions for  identification.  If other  sources were used, enter
 sufficient information  in the  available 12  spaces  to identify  the
 manufacturer and  the  solution  used.

 Use additional FORMs  II(PART 1)-IN  if  more  calibration sources were
 used.

 Under  "Initial Calibration True", enter the value  (in ug/L, to one
 decimal place) of the concentration of each analyte in the Initial
 Calibration Verification Solution.

Under  "Initial Calibration Found",  enter  the most recent value (in ug/L,
 to two decimal places),  of the  concentration of each analyte measured in
 the Initial Calibration  Verification Solution.

Under  "Initial Calibration %R",  enter  the value (to one decimal place)
of the percent recovery  computed according  to  the following equation:

  %R       =    	Found(ICV)	 x  1QO             (2  1}
                   True(ICV)

Where,  True(ICV)  is the  true concentration  of  the analyte in the Initial
Calibration Verification  Solution and  Found(ICV)  is the found
                              B-21                                ILM01.0

-------
concentration of the analyte in the Initial Calibration Verification
Solution.

The values used in equation 2.1 for True(ICV) and Found(ICV) must be
exactly those reported on this form.

Under "Continuing Calibration True", enter the value (in ug/L, to one
decimal place) of the concentration of each analyte in the Continuing
Calibration Verification Solution.

Under "Continuing Calibration Found", enter the value (in ug/L, to two
decimal places) of the concentration of each analyte measured in the
Continuing Calibration Verification Solution.

Note that the form contains two "Continuing Calibration Found" columns.
The column to the left must contain values for the first Continuing
Calibration Verification, and the column to the right must contain
values for the second Continuing Calibration Verification.  The column
to the right should be left blank if no second Continuing Calibration
Verification was performed.

If more than one FORM II(PART 1)-IN is required to report multiple
Continuing Calibration Verifications, then the column to the left on the
second form must contain values for the third Continuing Calibration
Verification, the column to the right must contain values for the fourth
Continuing Calibration Verification, and so on.

Under "Continuing Calibration %R", enter the value (to one decimal
place) of the percent recovery computed according to the following
equation:

   %R      =      Found(CCV)	 xlOQ              (22)
                   True(CCV)

where, True(CCV) is the true concentration of each analyte, and
Found(CCV) is the found concentration of the analyte in the Continuing
Calibration Verification Solution.

The values used in equation 2.2 for True(CCV) and Found(CCV) must be
exactly those reported on this form.

Note that the form contains two "Continuing Calibration %R" columns.
Entries to these columns must follow the sequence detailed above for
entries to the "Continuing Calibration Found" columns.

Under "M", enter the method used  or "NR", as explained in Part C.

If more  than  one wavelength is used to analyze an analyte, submit
additional FORMs II(PART 1)-IN as appropriate.

The  order of  reporting ICVs and CCVs for each analyte must follow the
temporal  order in which  the standards were run starting with the first
Form IIA  and  moving  from the left to the right continuing to the
following Form IIA's as  appropriate.  For instance, the first ICV for

                               B-22                                ILM01.0

-------
     all analytes must be reported on the  first Form IIA.   In a run where
     three CCVs were analyzed,  the first CCV must be reported in the left CCV
     column on the first Form IIA and the  second CCV must be reported in the
     right column of the same form.  The third CCV must be reported in the
     left CCV column of the second Form IIA.   On the second Form IIA, the ICV
     column and the right CCV column must  be left empty in this example.   In
     the previous example, if a second run for an analyte was needed, the ICV
     of that run must be reported on a third Form IIA and the CCVs follow in
     the same fashion as explained before.   In the case where two wavelengths
     are use used for an analyte, all ICV  and CCV results of one wavelength
     from all runs must be reported before proceeding to report the results
     of the second wavelength used.

E.    CRDL Standard for AA and ICP [FORM II(PART 2)-IN]

     This form is used to report analyte recoveries from analyses of the  CRDL
     Standards for AA (CRA) and 2x the CRDL Standards for ICP (CRI).

     Complete the header information according to the instructions' in Part A
     and as follows.

     Enter the AA CRDL Standard Source (12 spaces maximum) and the ICP CRDL
     Standard Source (12 spaces maximum),  as explained in Part D.

     Under "CRDL Standard for AA  True," enter the value (in ug/L,  to one
     decimal place) of the concentration of each analyte in the CRDL Standard
     Source Solution that was analyzed.

     Under "CRDL Standard for AA  Found",  enter the value  (in ug/L,  to two
     decimal places) of the concentration  of each analyte  measured in the
     CRDL Standard Solution.

     Under "CRDL Standard for AA  %R",  enter the value  (to one decimal place)
     of the percent recovery computed according to the  following equation:

       %R    =         Found  CRDL Standard  for  AA     IQQ    /o 3-)
                       True CRDL Standard  for AA

     Under "CRDL Standard for ICP  Initial  True",  enter the value (to one
     decimal place) of the concentration of each analyte in the CRDL Standard
     Solution that was analyzed by ICP for analytical samples associated  with
     the SDG.  Concentration units are ug/L.

     Under "CRDL Standard for ICP  Initial  Found",  enter the value (to two
     decimal places) of the concentration  of each analyte  measured in the
     CRDL Standard Solution analyzed at  the beginning of each run.
     Concentration units are  ug/L.

     Under "CRDL Standard for ICP,  Initial  %R"  enter the value (to one
     decimal place) of the percent recovery computed according to the
     following equation:

       %R   =   CRDL Standard for ICP Initial Found   x -^QQ      ,~  ,-.
                     CRDL  Standard  for ICP True

                                   B-23                                ILM01.0

-------
Under "CRDL Standard for ICP  Final Found", enter the value  (in ug/L,  to
two decimal places) of the concentration of each analyte measured  in the
CRDL Standard Solution analyzed at the end of each run.

Under "CRDL Standard for ICP  Final %R", enter the value (to one decimal
place) of the percent recovery computed according to the following
equation:

    %R     -      CRDL Standard for ICP Final Found   x 10Q    /2  5\
                      CRDL Standard for ICP True

All %R values reported in equations 2.3,  2.4, and 2.5 must be calculated
using the exact true and found values reported on this form.

Note that for every initial solution reported there must be a final one.
However, the opposite is not true.  If a CRDL Standard for ICP (CRI) was
required to be analyzed in the middle of a run (to avoid exceeding the
8-hour limit),  it must be reported in the "Final Found" section of this
form.

If more CRI or CRA analyses were required or analyses were performed
using more than one wavelength per analyte, submit additional FORMs
II(PART 2)-IN as appropriate.

The order of reporting CRAs and CRIs for each analyte must follow  the
temporal order in which the standards were run starting with the first
Form IIB and continuing to the following Form IIB's as appropriate.  The
order of reporting CRA and CRI is independent with respect to each
other.  When multiple wavelengths are used for one analyte, all the
results of one wavelength must be reported before prodeeding to the next
wavelength.

Blanks [FORM III-IN]

This form is used to report analyte concentrations found in the Initial
Calibration Blank (ICB),  in Continuing Calibration Blanks (CCB),  and in
the Preparation Blank (PB).

Complete the header information according to the  instructions in Part A
and as follows.

Enter "SOIL" or "WATER" as appropriate as the matrix of the Preparation
Blank.  No abbreviations  or other matrix descriptors may be used.

According to the matrix specified for the Preparation Blank,  enter
"UG/L" (for water) or "MG/KG"  (for soil)  as the Preparation Blank
concentration units.

Under "Initial Calib.  Blank",  enter the concentration (in ug/L, to one
decimal place)  of each analyte in the most recent Initial Calibration
Blank.

Under the "C" qualifier field, for any analyte enter "B" if the absolute
value of the analyte concentration is less than the CRDL but greater

                              B-24                                ILM01.0

-------
     than or equal to the IDL.  Enter "U" if the absolute value of the
     analyte in the blank is less than the IDL.

     Under "Continuing Calibration Blank I", enter the concentration (in
     ug/L, to one decimal place) of each analyte detected in the first
     required Continuing Calibration Blank (CCB) analyzed after the Initial
     Calibration Blank.   Enter any appropriate qualifier, as explained for
     the "Initial Calibration Blank," to the "C" qualifier column immediately
     following the "Continuing Calibration Blank 1" column.

     If only one Continuing Calibration Blank was analyzed,  then leave the
     columns labeled "2" and "3" blank.   If up to three CCB's were analyzed,
     complete the columns labeled "2" and "3", in accordance with the
     instructions for the "Continuing Calibration Blank 1" column.  If more
     than three Continuing Calibration Blanks were analyzed, then complete
     additional FORMs III-IN as appropriate.

     Under "Preparation Blank", enter the concentration in ug/L (to three
     decimal places)  for a water blank or in mg/Kg (to three decimal places)
     for a soil blank,  of each analyte in the Preparation Blank.  Enter any
     appropriate qualifier,  as explained for the "Initial Calibration Blank,"
     to the "C" qualifier column immediately following the "Preparation
     Blank" column.

     For all blanks,  enter the concentration of each analyte (positive or
     negative)  measured above the IDL or below the negative  value of the IDL.

     Under "M", enter the method used, as explained in Part  C.

     If more than one wavelength is used to analyze an analyte,  submit
     additional FORMs III-IN as appropriate.

     The order  of reporting ICBs and CCBs for each analyte must follow the
     temporal order  in which the blanks  were run starting with the first Form
     HI and moving  from left to right and continuing to the following Form
     IIIls as explained in Part D.   When multiple wavelengths are used for
     the analysis of one analyte,  all the results of one wavelength must be
     reported before  proceeding to  the next wavelength.

G.    ICP Interference Check Sample  [FORM IV-IN]

     This form  is used to report Interference Check Sample (ICS) results for
     each ICP instrument used in Sample  Delivery Group analyses.

     Complete the header information according to the instructions in Part A
     and as follows:

     For "ICP ID Number",  enter an  identifier that uniquely  identifies  a
     specific instrument within the Contractor laboratory.   No  two ICP
     instruments within a laboratory may have the same ICP ID Number.

     Enter "ICS Source"  (12  spaces  maximum)  as explained in  Part D.   For EPA
     solutions  include  in the source name a number identifying  it  (e.g.,  EPA-
     LV87).

                                   B-25                                ILM01.0

-------
Under "True Sol. A", enter Che true concentration (in ug/L, to the
nearest whole number) of each analyte present in Solution A.

Under "True Sol. AB", enter the true concentration (in ug/L, to the
nearest whole number) of each analyte present in Solution AB.

Under "Initial Found Sol. A", enter the concentration (in ug/L, to the
nearest whole number) of each analyte found in the initial analysis of
Solution A as required in Exhibit E.

Under "Initial Found Sol. AB", enter the concentration (in ug/L, to one
decimal place) of each analyte in the initial analysis of Solution AB as
required in Exhibit E.

Under "Initial Found %R", enter the value (to one decimal place) of the
percent recovery computed for true solution AB greater than zero
according to the following equation:

    %R     =         Initial Found Solution AB     -,QQ          (2 61
                         True Solution AB

Leave the field blank if true solution AB equals zero.

Under "Final Found Sol. A", enter the concentration (in ug/L, to the
nearest whole number) of each analyte found in the final analysis of
Solution A as required in Exhibit E.

Under "Final Found Sol. AB", enter the concentration (in ug/L, to one
decimal place) of each analyte found in the final analysis of Solution
AB as required in Exhibit E.

For All Found values of solutions A and AB, enter the concentration
(positive, negative, or zero) of each analyte at each wavelength used
for analysis by ICP.

Under "Final Found %R" , enter the value (to one decimal place) of the
percent recovery computed according to the following equation:

    %R     =         Final  Found Solution AB      x -^QQ          (27)
                         True Solution AB

All %R values reported must be calculated using the exact true and found
values reported on this form.

Note  that for every  initial solution reported there must be a final one.
However, the opposite is not true.  If an ICS was required to be
analyzed in the middle of a run (to avoid exceeding the 8-hour limit),
it must be reported  in the "Final Found" section of this form.

If more ICS analyses were required, submit additional FORMs IV-IN as
appropriate.

The order of reporting ICSs for each analyte must follow the temporal
order in which  the standards were run starting with the first Form IV

                              B-26                                ILM01.0

-------
     and continuing to the following Form IVs as appropriate.  When multiple
     wavelengths are used for one analyte,  all the results of one wavelength
     must be reported before prodeeding to  the next wavelength.

H.   Spike Sample Recovery [FORM V(PART 1)-IN]

     This form is used to report results for the pre-digest spike.

     Complete the header information according to the  instructions  in Part A
     and as follows.

     Indicate the appropriate matrix,  level and concentration units (ug/L for
     water and mg/Kg dry weight for soil) as explained in Parts  A and C.

     For "%Solids for Sample," enter the percent solids (as explained in Part
     C)  for the original sample of the EPA  Sample Number reported on the
     form.   Note that this number must equal the one reported on Form I  for
     that sample.

     In  the "EPA Sample No." box,  enter the EPA Sample Number (7 places
     maximum) of the sample from which the  spike results on this form were
     obtained.   The number must be centered in the box.

     Under "Control Limit %R", enter "75-125"  if the spike  added value was
     greater than or equal to one-fourth of the sample result value.   If not,
     leave the field empty.

     Under "Spiked Sample Result (SSR)",  enter the measured value (to four
     decimal places),  in appropriate units,  for each relevant analyte in the
     matrix spike  sample.   Enter any appropriate qualifier,  as explained in
     Part C,  to the "C" qualifier  column immediately following the  "Spiked
     Sample Result (SSR)"  column.

     Under "Sample Result  (SR)",  enter the  measured value  (to four  decimal
     places)  for each  required analyte in the  sample (reported in the EPA
     Sample No.  box)  on which the  matrix spike was performed.  Enter  any
     appropriate qualifier,  as explained in Part C,  to the  "C" qualifier
     column immediately following  the  "Sample  Result (SR)"  column.

     Under "Spike  Added (SA)",  enter the value (to two decimal places) for
     the concentration of  each analyte added to the  sample.   The  same
     concentration units must be used  for spiked sample  results,  unspiked
     (original  sample)  results,  and  spike added sample  results.   If the
     "spike added"  concentration  is  specified  in the contract, the  value
     added  and  reported must  be  that specific  concentration  in appropriate
     units,  corrected  for  spiked sample  weight and % solids  (soils) or spiked
     sample volume  (waters).

     Under  "%R", enter  the  value  (to one decimal place)  of  the percent
     recovery for  all  spiked  analytes  computed according to  the following
     equation:


              %R    -      	(SSR -   SR)      x  100                (2 8)
                                  SA

                                  B-27                                ILM01.0

-------
 %R must be reported, whether  it  is negative, positive or zero.

 The values for SSR,  SR,  and SA must be exactly those reported  on  this
 form.  A value of zero must be used in calculations for SSR  or SR if the
 analyte value is less than the IDL.

 Under  "Q", enter "N" if  the Spike Recovery  (%R) is out of the  control
 limits (75-125) and  the  Spike Added (SA) is greater than or  equal to
 one-fourth of the Sample Result  (SR).

 Under  "M" , enter the method used (as explained in Part C) or enter "NR"
 if the analyte is not required in the spike.

 If different samples were used for spike sample analysis of  different
 analytes, additional FORMs V(PART 1)-IN must be submitted for  each
 sample as appropriate.

 Post Digest Spike Sample Recovery [FORM V(PART 2)-IN]

 This form is used to report results for the post-digest spike  recovery
 which is based upon  the addition of a known quantity of analyte to an
 aliquot of the digested sample.

 Complete the header  information according to the instructions  in  Part A
 and as follows.

 In the "EPA Sample No." box,  enter the EPA Sample Number (7  spaces
 maximum) of the sample from which the spike results on this  form  were
 obtained.  The number must be centered in the box.

 The "Control Limit %R" and "Q" fields must be left blank until  limits
 are established by EPA.   At that time,  the Contractor will be  informed
 how to complete these fields.

 Under "Spiked Sample Result (SSR)",  enter the measured value (in  ug/L,
 to two decimal places) for each analyte in the post-digest spike  sample.
 Enter any appropriate qualifier,  as explained in Part C,  to  the "C"
 qualifier column immediately following the "Spiked Sample Result  (SSR)"
 column.

Under "Sample Result (SR)",  enter the measured value (in ug/L,   to  two
 decimal places)  for  the concentration of each analyte in the sample
 (reported in the EPA Sample No.  box)  on which the spike was performed.
 Enter any appropriate qualifier,  as explained in Part C,  to  the "C"
 qualifier column immediately following the "Sample  Result (SR)" column.

Under "Spike Added (SA)", enter the value (in ug/L,  to one decimal
 place) for each analyte added to the sample.  The same concentration
 units must be used for spiked sample results,  unspiked (original  sample)
 results,  and spike added sample  results.   If the spike added
 concentration is specified in the contract,  the value added and reported
 must be that specific concentration in appropriate  units.
                              B-28                                ILM01.0

-------
Under "%R",  enter the value (to one decimal place) of the percent
recovery for all spiked analytes computed according to Equation 2.8 in
Part H, preceding.

%R must be reported, whether it is negative, positive or zero.

The values for SSR, SR, and SA must be exactly those reported on this
form.  A value of zero must be substituted for SSR or SR if the analyte
value is less than the IDL.

Under "M", enter the method used as explained in Part C, or enter "NR"
if the spike was not required.

If different samples were used for spike sample analysis of different
analytes, additional FORMS V(PART 1)-IN must be submitted.

Duplicates [FORM VI-IN]

The duplicates form is used to report results of duplicate analyses.
Duplicate analyses are required for % solids values and all analyte
results.

Complete the header information according to the instructions in Part A
and as follows.

Indicate the appropriate matrix, level and concentration units (ug/L for
water and mg/Kg dry weight for soil) as explained in Parts A and C.

For "% Solids for Sample," enter to percent solids (as explained in Part
C) for the original sample of the EPA Sample Number reported on the
form.  Note that this number must equal the one reported on Form I for
that sample.

For "% Solids for Duplicate," enter the percent solids (as explained in
Part C) for the duplicate sample of the EPA Sample Number reported on
the,form.

In the "EPA Sample No." box, enter the EPA Sample Number  (7 spaces
maximum) of the sample from which the duplicate sample results on this
form were obtained.  The number must be centered in the box.

Under "Control'Limit",  enter the CRDL (in appropriate units, ug/L for
water or mg/kg dry weight basis compared to the original sample weight
and percent solids) for the analyte if the sample or duplicate values
were less than 5x CRDL and greater than or equal to the CRDL.   If the
sample and duplicate values were less than the CRDL or greater than or
equal to 5x CRDL, leave the field empty.

Under Sample (S), enter the original measured value (to four decimal
places) for the concentration of each analyte in the sample (reported in
the EPA Sample No. box) on which a Duplicate analysis was performed.
Concentration units are those specified on the form.  Enter any
appropriate qualifier,  as explained in Part C, to the "C" qualifier
column immediately following the "Sample (S)" column.

                              B-29                                ILM01.0

-------
     Under Duplicate (D),  enter the measured value  (to four decimal places)
     for each analyte in  the Duplicate sample.   Concentration units are those
     specified on the form.   Enter any appropriate qualifier,  as explained in
     Part C,  to the "C" qualifier column immediately following the "Duplicate
     (D)" column.

     For solid samples, the  concentration of the original sample must be
     computed using the weight and % solids  of  the original sample.  The
     concentration of the  duplicate sample must be computed using the weight
     of the duplicate sample,  but the % solids  of the original sample.

     Under RPD,  enter the  absolute value (to one decimal place) of the  •
     Relative Percent Difference for all analytes detected above the IDL in
     either the sample or  the  duplicate, computed according to the following
     equation:

                     RPD  -         *  S ' D *      x   100         (2.9)
                                   (S + D)/2

     The values for S and  D  must be exactly  those reported on this form.  A
     value of zero must be substituted for S or D if the analyte
     concentration is less than the IDL in either one.   If the analyte
     concentration is less than the IDL in both S and D, leave the RPD field
     empty.

     Under "Q" ,  enter "*"  if the duplicate analysis for the analyte is out of
     control.  If both sample  and duplicate  values are greater than or equal
     to 5x CRDL,  then the  RPD  must be less than or equal to 20% to be in
     control.  If either  sample or duplicate values are less than 5x CRDL,
     then the absolute difference between the two values must be less than
     the CRDL to be in control.   If both values are below the CRDL, then no
     control limit is applicable.

     Under "M",  enter method used as explained  in Part C.

K.   Laboratory Control Sample [FORM VII-IN]

     This form is used to  report results for the solid and aqueous Laboratory
     Control Samples.

     Complete the header  information according  to the instructions in Part A
     and as follows.

     For the Solid LCS Source (12 spaces maximum), enter the appropriate EPA
     sample number if the  EPA provided standard was used.  Substitute an
     appropriate number provided by the EPA  for LCS solutions prepared in the
     future.   If other sources were used, complete as explained in Part D.
     For the Aqueous LCS  Source, enter the source name (12 spaces maximum) as
     explained in Part D.

     Under "Aqueous True", enter the value (in  ug/L, to one decimal place) of
     the concentration of each analyte in the Aqueous LCS Standard Source.
                                   B-30                                ILM01.0

-------
     Under "Aqueous Found", enter the measured concentration  (in ug/L, to  two
     decimal places) of each analyte found in the Aqueous LCS solution.

     Under "Aqueous %R" ,  enter the value of the percent recovery (to one
     decimal place) computed according to the following equation:

        %R    -      Aqueous LCS  Found    ^ IQQ                   (2 1Q)
                     Aqueous LCS  True

     Under "Solid True",  enter the value (in mg/Kg, to one decimal place)  of
     the concentration of each analyte in the Solid LCS Source.

     Under "Solid Found", enter  the measured value (in mg/Kg, to one decimal
     place) of each analyte found in the Solid LCS solution.

     Under "C",  enter "B" or "U" or leave empty, to describe the found value
     of the solid LCS as  explained in Part C.

     Under "Limits", er^er the lower limit (in mg/Kg, to one decimal place)
     in the left column,  and the upper limit (in mg/Kg, to one decimal place)
     in the right column, for each analyte in the Solid LCS Source solution.

     Under "Solid %R",  enter the value of the  percent recovery (to one
     decimal place) computed according to the  following equation:

        %R    -      Solid LCS Found      ^ IQQ                   (2
                      Solid LCS True

     The values  for true  and found aqueous and solid LCS's used in equations
     2.10 and 2.11 must be exactly those reported on this form.   If the
     analyte concentration is less than the IDL, a value of zero must be
     substituted for the  solid LCS found.

     Submit additional FORMs VII-IN as appropriate, if more than one aqueous
     LCS or solid LCS was required.

L.   Standard Addition Results [FORM VIII-IN]

     This form is used to report  the results of samples analyzed using the
     Method of Standard Additions (MSA)  for Furnace AA analysis.

     Complete the header  information according to the instructions  in Part A.

     Under "EPA  Sample  No.",  enter the EPA Sample Numbers (7 spaces maximum)
     of all analytical  samples analyzed using  the MSA.   This includes  reruns
     by MSA (if,the first MSA was out of control)  as  explained in Exhibit E.

     Note that only field samples and duplicates may be reported on this
     form,  thus  the EPA Sample Number usually  has no  suffix or a "D."

     A maximum of 32 samples can be  entered on this form.   If additional
     samples required MSA,  submit additional FORMs VIII-IN.   Samples  must be
                                   B-31                                ILM01.0

-------
listed in alphanumeric order per analyte, continuing to the next FORM
VIII-IN if applicable.

Under "An", enter the chemical symbol (2 spaces maximum) for each
analyte for which MSA was required for each sample listed.  The analytes
must be in alphabetic listing of the chemical symbols.

Results for different samples for each analyte must be reported
sequentially,  with the analytes ordered according to the alphabetic
listing of their chemical symbols.  For instance, results for As
(arsenic) in samples MAA110, MAAlll, and MAA112 would be reported in
sequence, followed by the result for Pb (lead) in MAA110 etc.

Under "0 ADD ABS", enter the measured value in absorbance units (to
three decimal places) for the analyte before any addition is performed/

Under "1 ADD CON", enter the final concentration in ug/L (to two decimal
places) of the analyte (excluding sample contribution) after the first
addition to the sample analyzed by MSA.

Under "1 ADD ABS", enter the measured value (in the same units and
decimal places as "0 ADD ABS") of the sample solution spiked with the
first addition.

Under "2 ADD CON", enter the final concentration in ug/L (to two decimal
places) of the analyte (excluding sample contribution) after the second
addition to the sample analyzed by MSA.

Under "2 ADD ABS", enter the measured value (in the same units and
decimal places as "0 ADD ABS") of the sample solution spiked with the
second addition.

Under "3 ADD CON", enter the final concentration in ug/L (to two decimal
places) of the analyte (excluding sample contribution) after the third
addition to the sample analyzed by MSA.

Under "3 ADD ABS", enter the measured value (in the same units and
decimal places as "0 ADD ABS") of the sample solution spiked with the
third addition.

Note that "0 ADD ABS", "1 ADD ABS" , "2 ADD ABS", and "3 ADD ABS" must
have the same dilution factor.

Under "Final Cone.", enter the final analyte concentration (in ug/L, to
one decimal place) in the sample  as determined by MSA computed
according to the  following formula:

Final Cone.    =    -  (x-intercept)                          (2.12)

Note that the final concentration of an analyte does not have to equal
the value for that analyte which is reported on FORM I-IN for that
s amp1e.
                              B-32                                ILM01.0

-------
     Under "r", enter the correlation coefficient (to four decimal places)
     that is obtained for the least squares regression line representing the
     following points (x,y):(0.0, "0 ADD ABS"), ("1 ADD CON", "1 ADD ABS"),
     ("2 ADD CON", "2 ADD ABS"), ("3 ADD CON", "3 ADD ABS").

     Note that the correlation coefficient must be calculated using the
     ordinary least squares linear regression (unweighted) according to the
     following formula:

                            N  £  ^i  -  I xi X y±
               r	_	    (2.13)
                   (N £ Xi2 - ( £ x^2]14 [N I Yi2 - ( I yi)2]h

     Under "Q", enter "+" if r is less than 0.995.  If r is greater than or
     equal to 0.995, then leave the field empty.

M.   TCP Serial Dilution [FORM IX-IN]

     This form is used to report results for ICP serial dilution.

     Complete the header information according to the instructions in Part A
     and as follows.

     In the "EPA Sample No." box, enter the EPA Sample Number (7 places
     maximum) of the sample for which serial dilution analysis results on
     this form were obtained.  The number must be centered in the box.

     Under "Initial Sample Result (I)", enter the measured value (in ug/L, to
     two decimal places) for each ICP analyte in the undiluted sample (for
     the EPA sample number reported on this form).  Enter any appropriate
     qualifier, as explained in Part C, to the "C" qualifier column
     immediately following the "Initial Sample Result (I)" column.

     Note that the Initial Sample Concentration for an analyte does not have
     to equal the value for that analyte reported on FORM I - IN for that
     sample.  It is the value of the analyte concentration (uncorrected for
     dilution) that is within the linear range of the instrument.

     Under "Serial Dilution Result (S)", enter the measured concentration
     value (in ug/L, to two decimal places) for each ICP analyte in the
     diluted sample.  The value must be adjusted for that dilution.  Enter
     any appropriate qualifier, as explained in Part B, to the "C" qualifier
     column immediately following the "Serial Dilution Result (S)" column.

     Note that the Serial Dilution Result (S) is obtained by multiplying by
     five the instrument measured value (in ug/L) of the serially diluted
     sample and that the "C" qualifier for the serial dilution must be
     established based on the serial dilution result before correcting it for
     the dilution regardless of the value reported on the form.

     Under "% Difference", enter the absolute value (to one decimal place) of
     the percent difference in concentration of required analytes,  between
     the original sample and the diluted sample (adjusted for dilution)
     according to the following formula:

                                   B-33                                ILM01.0

-------
       % Difference   -         \  I  - S  \   x 100                 (2.14)


     The values for I and S used to calculate % Difference in equation 2.14
     must be exactly those reported on this form.  A value of zero must be
     substituted for S if the analyte concentration is less than the IDL.  If
     the analyte concentration in (I) is less than the IDL, concentration
     leave "% Difference" field empty.

     Under "Q",  enter "E" if the % Difference is greater than 10% and the
     original sample concentration (reported on FORM I-IN) is greater than
     50x the IDL reported on FORM X-IN.

     Under "M",  enter the method of analysis for each analyte as explained in
     Part C.

N.   Instrument Detection Limits (Quarterly) [FORM X-IN]

     This form documents the Instrument  Detection Limits  for each instrument
     that the laboratory used to obtain  data for the Sample Delivery Group.
     Only the instrument and wavelengths used to generate data for the SDG
     must be included.

     Although the Instrument Detection Limits (IDLs) are  determined quarterly
     (every three calendar months) a copy of the quarterly instrument
     detection limits must be included with each SDG data package on FORM(s)
     X-IN.

     Complete the header information according to the instructions in Part A
     and as follows.

     Enter the date (formatted MM/DD/YY) on which the IDL values were
     obtained (or became effective).

     Enter ICP ID Number, Flame AA ID Number,  and Furnace AA ID Number (12
     spaces maximum each).   These ID Numbers are used to  uniquely identify
     each instrument that the laboratory uses to do CLP work.

     Enter the Mercury instrument ID number in the Flame  AA ID Number field.

     Under "Wavelength",  enter the wavelength in nanometers (to two decimal
     places) for each analyte for which  an Instrument Detection Limit (IDL)
     has  been established and is listed  in the IDL column.   If more than one
     wavelength is used for an analyte,  use other FORMs X-IN as appropriate
     to report the Instrument Detection  Limit.

     Under "Background",  enter the type  of background correction used to
     obtain Furnace AA data.   Enter "BS" for Smith Hieftje, "BD" for
     Deuterium Arc, or "BZ" for Zeeman background correction.

     Contract Required Detection Limits  (in ug/L) as established in Exhibit
     C, must appear in the column headed "CRDL".
                                   B-34                                ILM01.0

-------
     Under "IDL",  enter the Instrument Detection Limit (ug/L, to one decimal
     place) as determined by the laboratory for each analyte analyzed by the
     instrument for which the ID Number is listed on this form.  Except for
     Mercury, the  instrument detection limit must be rounded to a whole
     number.

     Under "M", enter the method of analysis used to determine the instrument
     detection limit for each wavelength used.   Use appropriate codes as
     explained in  Part C.

     Use additional FORMs X-IN if more instruments and wavelengths are used.
     Note that the date on this form must not exceed the analysis dates in
     the SDG  data  package or precede them by more than three months.

     Use the  Comments section to indicate alternative wavelengths and the
     conditions under which they are used.

0.   ICP Interelement Correction Factors (Annually) [FORM XI(PART 1)-IN]

     This form documents for each ICP instrument tha i: rerelement correction
     factors  applied by the Contractor laboratory to obtain data for  the
     Sample Delivery Group.

     Although the  correction factors are determined annually (every twelve
     calendar months),  a copy of the results of the annual interelement
     correction factors must be included with each SDG data package on FORM
     XI(PART  1)-IN.

     Complete the  header information according to instructions  in Part A and
     as follows.

     Enter the ICP ID Number (12 spaces maximum),  which is a unique number
     designated by the laboratory to identify each ICP instrument used to
     produce  data  in the SDG package.   If more  than one ICP instrument is
     used,  submit  additional FORMs XI(PART 1)-IN as appropriate.

     Report the date (formatted as MM/DD/YY)  on which these correction
     factors  were  determined for use.    This  date must not exceed the ICP
     analysis dates  in the SDG data package or  precede them by  more than
     twelve calendar months.

     Under "Wavelength",  list the wavelength  in nanometers (to  two  decimal
     places)  used  for each ICP analyte.   If more than one  wavelength  is  used,
     submit additional  FORMs  XI(PART 1)-IN as  appropriate.

     Under "Al", "Ca",  "Fe",  "Mg",  enter the  correction factor  (negative,
     positive or zero,  to seven decimal places,  10  spaces  maximum)  for  each
     ICP analyte.   If correction factors for  another analyte are  applied,  use
     the empty column and list the analyte's  chemical  symbol in  the blank
     two-space header field provided for that  column.

     If corrections  are not applied for an analyte,  a  zero muse be  entered
     for that analyte to indicate that the corrections  were  determined  to  be
                                   B-35                               ILM01.0

-------
     zero.   If correction factors are applied for more than one additional
     analyte,  use FORM XI(PART 2)-IN.

P.   ICP Interelement Correction Factors (Annually)  [FORM  XI(PART 2)-IN]

     This form is used if correction factors for analytes other than Al,  Ca,
     Fe,  Mg,  and one more analyte of the Contractor's choice,  were applied to
     the analytes analyzed by ICP.   Complete this form as for  FORM XI(PART
     1)-IN by listing the chemical symbol for additional analytes in the
     heading of the empty columns in the two-space fields provided.

     Columns of correction factors for additional analytes must be entered
     left to right starting on FORM XI(PART 1)-IN and proceeding to FORM
     XI(PART 2)-IN, according to the alphabetic order of their chemical
     symbols.   Note that correction factors for Al,  Ca,  Fe,  and Mg are all
     required and are to be listed first (as they appear on FORM XI(PART  1)-
     IN).

Q.   ICP Linear Ranges (Quarterly)  [FORM XII-IN]

     This form documents the quarterly linear range  analysis for each ICP
     instrument that the laboratory used to obtain data for the SDG.

     Complete the header information according to the instructions in Part A
     and as follows.

     Enter the ICP ID Number (12 spaces maximum),  which is a unique number
     designated by the Contractor to identify each ICP instrument used to
     produce data for the SDG.  If more than one ICP instrument is used,
     submit additional FORMs XII-IN as appropriate.

     Report the date (formatted as MM/DD/YY) on which these linear ranges
     were determined for use.  This date must not exceed the dates of
     analysis by ICP in the SDG data package and must not precede the
     analysis dates by more than three calendar months.

     Under "Integ. Time (Sec.)," enter the integration time (in seconds to
     two decimal places) used for each measurement taken from the ICP
     ins trument.

     Under "Concentration", enter the concentration  (in ug/L)  that is the
     upper limit of the ICP instrument linear range  as determined in Exhibit
     E.   Any measurement in the SDG data package at  or below this
     concentration is within the linear range.  Any  measurement above it  is
     out of the linear range, and thus, is an estimated value  and must be
     diluted into the linear range.
               ;
     Under "M", enter the method of analysis for each analyte  as explained in
     Part C.

     If more instruments or analyte wavelengths are  used, submit additional
     FORMs XII-IN as appropriate.
                                   B-36                                ILM01.0

-------
R.   Preparation Log [Form XIII-IN]

     This Form is used to report the preparation run log.

     All field samples and all quality control preparations (including
     duplicates, matrix spikes, LCS's, PB's and repreparations) associated
     with the SDG must be reported on Form XIII.

     Submit one Form XIII per batch, per method,  if no more than thirty-two
     preparations, including quality control preparations,  were performed.
     If more than thirty-two preparations per batch, per method,  were
     performed, then submit additional copies of Form XIII  as appropriate.
     Submit a separate Form XIII for each batch.

     The order in which the Preparation Logs are submitted  is very important.
     Form XIII must be organized by method, by batch.  Later batches within a
     method must follow earlier ones.  Each batch must start on a separate
     Form XIII.

     Complete the header information according to the instructions in Pajt A,
     and as follows:

     For "Method", enter the method of analysis (two characters maximum) for
     which the preparations listed on the Form were made.   Use appropriate
     method codes as specified in Part C.

     Under "EPA Sample No.", enter the EPA Sample Number of each sample in
     the SDG, and of all other preparations such as duplicates, matrix
     spikes, LCSs, PBs, and repreparations (all formatted according to Table
     1).  All EPA Sample Numbers must be listed in ascending alphanumeric
     order, continuing to the next Form XIII if applicable.

     Under "Preparation Date", enter the date (formatted MM/DD/YY) on which
     each sample was prepared for analysis by the method indicated in the
     header section of the Form.

     Note that the date never changes on a single Form XIII because the form
     must be submitted per batch.

     Under "Weight",  enter the wet weight (in grams, to two decimal places)
     of each soil sample prepared for analysis by the method indicated in the
     header section of the Form.  If the sample matrix is water,  then leave
     the field empty.

     Under "Volume",  enter the final volume (in mL, to the  nearest whole
     number) of the preparation for each sample prepared for analysis by the
     method indicated  in the header section of the Form.  This field must
     have a value for  each sample listed.
                                   B-37                                ILM01.0

-------
S.    Analysis Run Log [Form XIV-IN]

     This  Form is used Co report the  sample analysis run  log.

     A  run is defined as  the totality of  analyses performed by an  instrument
     throughout the  sequence initiated by, and  including,  the first  SOW-
     required calibration standard and terminated by, and including,  the
     continuing calibration verification  and blank  following the last  SOW-
     required analytical  sample.

     All field samples and all  quality control  analyses  (including
     calibration standards,  ICVs, CCVs, ICBs, CCBs, CRAs,  CRIs, ICSs,  LRSs,
     LCSs,  PBs,  duplicates,  serial dilutions, pre-digestion spikes,  post-
     digestion spikes,  analytical spikes, and each  addition analyzed for  the
     method of standard addition determination) associated with the  SDG must
     be reported on  Form  XIV.   The run must be  continuous and inclusive of
     all analyses performed on  the particular instrument  during the  run.

     Submit one Form XIV  per run if no more than thirty-two (32) analyses,
     including instrument calibration, were analyzed in the run.   If more
     than  thirty-two analyses were performed in the run,  submit additional
     Forms  XIV as appropriate.

     The order in which the Analysis  Run  Logs are submitted is very
     important.   Form XIV must  be organized by  method, by run.  Later  runs
     within a method must follow earlier  ones.  Each analytical run  must
     start:  on a separate  Form XIV.  Therefore,  instrument calibration  must be
     the first entry on the form for  each new run.  In addition, the run  is
     considered to have ended if it is interrupted  for any reason, including
     termination for failing QC parameters.

     Complete the header  information  according  to the instructions in  Part A,
     and as follows:

     For "Instrument ID Number",  enter the instrument ID  number, (12 spaces
     maximum) , which must be an identifier designated by  the laboratory to
     uniquely identify each instrument used to  produce data which  are
     required to be  reported in the SDG deliverable.  If  more than one
     instrument is used,  submit additional Forms XIV as appropriate.

     For "Method", enter  the method code  (two characters  maximum)  according
     to the specifications in Part C.

     For "Start Date",  enter the date (formatted MM/DD/YY) on which  the
     analysis run was started.

     For "End Date",  Enter the  date (formatted  MM/DD/YY)  on which  the
     analysis run was ended.

     Under  "EPA Sample No.",  enter the EPA sample number  of each analysis,
     including all QC operations  applicable to  the  SDG (formatted  according
     to Table 1). All EPA Sample Numbers must  be listed  in increasing
     temporal (date  and time) order of analysis, continuing t:o the next Form
     XIV for the instrument run if applicable.  The analysis date  and  time of

                                  B-38                                ILM01.0

-------
 other analyses  not associated with the SDG,  but analyzed by the
 instrument in the  reported analytical run,  must be reported.   Those
 analyses must be  identified with the EPA Sample No.  of "ZZZZZZ".

 Under "D/F",  enter the dilution factor (to  two decimal places) by which
 the final digestate or distillate needed to be diluted for each analysis
 to  be performed.   The  dilution factor does  not include the dilution
 inherent in  the preparation as specified by the preparation procedures
 in  Exhibit D.

 The dilution factor is required for all  entries on Form XIV.

 Note that for a particular sample a dilution factor  of "1" must be
 entered if the  digestate  or distillate were  analyzed without  adding any
 further volume  of  dilutant or any other  solutions  to the "Volume"  or an
 aliquot of the  "Volume" listed on Form XIII  for that sample.

 For EPA supplied solutions such as  ICVs,  ICSs,  and LCSs,  a. dilution
 factor must be  entered if the supplied solution had  to be diluted  to a
 dilution different from that  specified by the  instructions provided with
 the solution.   The dilution factor  reported  in such  a case must be that
 which would make the reported true  values on the appropriate   form for
 the solution  equal those  that were  supplied  with the solution by  the
 EPA.   For instance, ICV-2(0887)  has a  true value of  104.0 ug/L at  a 20
 fold dilution.  If the solution is  prepared  at a 40  fold dilution,  a
 dilution factor of "2" must be  entered on Form XIV and the uncorrected
 instrument reading is  compared to a true  value of  52 ug/L.  In this
 example,  Form II will  have a  true value  of 104.0 regardless of the
 dilution used.  The found value  for the  ICV  must be  corrected for  the
 dilution listed on Form XIV using the  following formula:

 Found value on  Form II -  Instrument  readout  in ug/L  x   D/F

 Under "Time", enter the time,  (in military format  -  HHMM),  at which each
 analysis  was performed.   If an  auto sampler  is  used  with  equal  analysis
 time  and  intervals between analyses, then only the start  time  of the run
 (the  time of analysis  of  the  first  calibration standard)  and  end time  of
 the  run (the time  of analysis  of the final CCV  or  CCB,  which  ever  is
 later)  need to  be  reported.

 Under  "%  R", enter the percent recovery  (to  one  decimal place)  for  each
 Furnace AA analytical  spike analyzed.  If the  analytical  spike was
 performed on more  than one  analyte, use additional Forms XIV  as
 appropriate.   Leave the "%  R" field empty if the analysis  reported  is
 not  for an analytical  spike.  %R must be  recorded  even  if  the result is
 not used.

A %R value of "-9999.9" must be entered for  the analytical spike if
 either  the sample  or analytical  results is greater than the calibration
 range of  the instrument.

Under  "Analytes",   enter "X" in the  column of the designated analyte to
 indicate  that the  analyte value was used  from  the reported analysis to
                              B-39                                ILM01.0

-------
     report.data in Che SDG.   Leave Che column empty for each analyce if the
     analysis was not used to report the particular analyte.

     Entering "X" appropriately is very important.   The "X" is used to link
     the samples with their related QC.  It also links the dilution factor
     with the appropriate result reported on Forms  I-IX.  For each analyte
     result reported on any of the Forms I-IX, there must be  one,  and only
     one,  properly identified entry on Form XIV for which an  "X" is entered
     in the column for that analyte.

T.   Sample Log-In Sheet [Form DC-1]

      This form is used to document the receipt and inspection of samples and
      containers.  One original of Form DC-1 is required for each sample
      shipping container, e.g., cooler.  If the samples in a single sample
      shipping container must be assigned to more than one Sample Delivery
      Group, the original Form DC-1 shall be placed with the  deliverables for
      the Sample Delivery Group of the lowest Arabic number and a copy of
      Form DC-1 must be placed with the deliverables for the  other Sample
      Delivery Group(s).  The copies should be identified as  "copy(ies)," and
      the location of the original should be noted on the copies.

      Sign and date the airbill (if present).  Examine the shipping container
      and record the presence/absence of custody seals and their condition
      (i.e., intact, broken) in item 1 on Form DC-1.  Record the custody seal
      numbers in item 2.

      Open the container, remove the enclosed sample documentation, and
      record the presence/absence of chain-of-custody record(s), SMO forms
      (i.e., Traffic Reports, Packing Lists), and airbills or airbill
      stickers in items 3-5 on Form DC-1.  Specify if there is an airbill
      present or an airbill sticker in item 5 on Form DC-1.  Record the
      airbill or sticker number in item 6.

      Remove the samples from the shipping container(s), examine the samples
      and the sample tags (if present), and record the condition of the
      sample bottles (i.e., intact, broken, leaking) and presence or absence
      of sample tags in items 7 and 8 on Form DC-1.

      Review the sample shipping documents and complete the header
      information described in Part A.  Compare the information recorded on
      all the documents and samples and mark the appropriate answer in item 9
      on Form DC-1.

      If there are no problems observed during receipt, sign and date
      (include time) Form DC-1, the chain-of-custody record,  and Traffic
      Report, and write the sample numbers on Form DC-1.  Record the
      appropriate sample tags and  assigned laboratory numbers if applicable.
      The log-in date should be recorded at  the top of  Form DC-1 and the date
      and time of cooler receipt at the laboratory should be recorded in
      items  10 and 11.  Cross out  unused columns and spaces.

      If there are problems observed during  receipt, contact the SMO and
      document the contact as well  as  resolution of the problem on  a CLP

                                    B-40                                ILM01.0

-------
      Communication Log.  Following resolution, sign and date the forms as
      specified in the preceding paragraph and note, where appropriate, the
      resolution of the problem.

      Record the fraction designation (if appropriate) and the specific area
      designation (e.g., refrigerator number) in the Sample Transfer block
      located in the bottom left corner of Form DC-1.  Sign and date the
      sample transfer block.

U.   Document Inventory Sheet (Form DC-2)

      This form is used to record the inventory of the Complete SDG File
      (CSF) documents which are sent to the Region.

      Organize all EPA-CSF documents as described in Exhibit B, Section II
      and Section III.  Assemble the documents in the order specified on Form
      DC-2 and Section II and III, and stamp each page with the consecutive
      number.  (Do not number the DC-2 form).  Inventory the CSF by reviewing
      the document numbers and recording page numbers ranges in the column
      provided on the Form DC-2.  If there are no documents for a specific
      document type, enter an "NA" in the empty space.

      Certain laboratory specific documents related to the CSF may not fit
      into a clearly defined category.  The laboratory should review DC-2 to
      determine if it is most appropriate to place them under No. 29, 30, 31,
      or 32.  Category 32 should be used if there is no appropriate previous
      category.  These types of documents should be described or listed in
      the blanks under each appropriate category.
                                   B-41                                ILM01.0

-------
     SECTION IV




DATA REPORTING FORMS
    B-42                                ILM01.0

-------
Lab Name:

Lab Code:

SOW No.:
                U.S.  EPA - CLP

 COVER PAGE -  INORGANIC ANALYSES  DATA PACKAGE


	        Contract:  	

                           SAS No.:
Case No.:
                                      SDG No.
               EPA Sample No.
                                Lab  Sample ID.
Were ICP interelement corrections applied?

Were ICP background corrections applied?
     If yes-were raw data generated before
     application of background corrections?

Comments:
                                                  Yes/No

                                                  Yes/No

                                                  Yes/No
I certify that this data package is in compliance with the terms and
conditions of the contract, both technically and for completeness, for other
than the conditions detailed above.  Release of the data contained in this
hardcopy data package and in the computer-readable data submitted on
diskette has been authorized by the Laboratory Manager or the Manager's
designee, as verified by the following signature.
Signature:_

Date:
                            Name:

                           Title:
                                COVER PAGE - IN
                                                     3/90

-------
                               U.S. EPA - CLP
                                                             EPA SAMPLE NO.
Lab Name:



Lab Code:
                         INORGANIC ANALYSIS DATA SHEET
               Contract:
Case No.:
SAS No.:
SDG No.:
Matrix (soil/water):



Level (low/med):



% Solids:
                           Lab Sample ID:



                           Date Received:
            Concentration Units (ug/L or rag/kg dry weight):
Color Before:



Color After:



Comments:
CAS No.
7429-90-5
7440-36-0
7440-38-2
7440-39-3
7440-41-7
7440-43-9
7440-70-2
7440-47-3
7440-48-4
7440-50-8
7439-89-6
7439-92-1
7439-95-4
7439-96-5
7439-97-6
7440-02-0
7440-09-7
7782-49-2
7440-22-4
7440-23-5
7440-28-0
7440-62-2
7440-66-6


1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
| Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
(Iron
[Lead
(Magnesium
| Manganese
| Mercury
| Nickel
[Potassium
| Selenium
| Silver
| Sodium
(Thallium
| Vanadium
| Zinc
| Cyanide
1
Concentration

























                                                            M
       Clarity Before:



       Clarity After:
                 Texture:



                 Artifacts:
                                   FORM I  - IN
                                              3/90

-------
                               U.S. EPA - CLP
                                     2A
               INITIAL AND CONTINUING CALIBRATION VERIFICATION
Lab Name:

Lab Code:
Case No.:
Contract:

SAS No.:
Initial Calibration Source:

Continuing Calibration Source:
SDG No.:
                          Concentration Units:  ug/L
Analyte
Aluminum
Antimony
Arsenic 	
Barium
Beryllium
Cadmium 	
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide

Initial Calibration
True Found %R ( 1 )











































































Continuing Calibration
True Found %R(1) Found %R(1)




























































































































                                                                          M
(1)   Control Limits: Mercury 80-120; Other Metals 90-110; Cyanide 85-115
                             FORM II (PART 1)  - IN
                                              3/90

-------
                               U.S. EPA - CLP
                                     2B
                         CRDL STANDARD FOR AA AND ICP
Lab Name:

Lab Code:
Case No.:
Contract:

SAS No.:
SDG No.:
AA CRDL Standard Source:

ICP CRDL Standard Source:
                          Concentration Units:  ug/L
1
1
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
| Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
| Iron
|Lead
Magnesium
Manganese
| Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc

CRL.,., S
True
























tandard fc
Found
























r AA
%R
























                                      True
                      CRDL Standard for ICP
                     Initial              Final
                       Found    %R     Found    %R
                             FORM II (PART 2}  - IN
                                             3/90

-------
                               U.S. EPA - CLP

                                      3
                                   BLANKS
Lab Name:

Lab Code:
Case No.:
Contract:

SAS No.:
                                      SDG No.
Preparation Blank Matrix (soil/water):
Preparation Blank Concentration Units (ug/L or mg/kg):
1
1
1
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
| Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
| Iron
[Lead
|:iagnesium
Manganese
I Mercury
I Nickel
| Potassium
I Selenium
I Silver
| Sodium
[Thallium
Initial
Calib.,
Blank
(ug/L) C





















[ Vanadium |
Zinc
| Cyanide







—




—


~








Continuing Calibration
Blank (ug/L)
1 C 2 C 3 C




















































































~
~































































Prepa-
ration
Blank C






























—




—











M








—















                                 FORM III - IN
                                              3/90

-------
                               U.S. EPA - CLP
                        ICP INTERFERENCE  CHECK SAMPLE
Lab Name:

Lab Code:
ICP ID Number:
Case No.
Contract:

SAS No.:
                   ICS Source:
SDG No.:
                          Concentration Units:  ug/L
1
1
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
| Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
I Iron
[Lead
| Magnesium
| Manganese
| Mercury
| Nickel
| Potassium
| Selenium
| Silver
| Sodium
[Thallium
[Vanadium
| Zinc
1
Ti
Sol.
A
























rue
Sol.
AB
























                                  Initial  Found
                              Sol.     Sol.
                               A        AB      %R
                                   Final  Found
                               Sol.     Sol.
                                A        AB
                                  FORM IV - IN
                                              3/90

-------
                                U.S.  EPA  -  CLP
Lab Name:

Lab Code:
                                       5A
                             SPIKE SAMPLE RECOVERY
                                       Contract:
                                     EPA SAMPLE NO.
Case No.:
SAS No.:
SDG No.:
Matrix  (soil/water):

% Solids for Sample:
                              Level  (low/med):
            Concentration Units (ug/L or mg/kg dry weight):
I
1
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
I Cadmium
| Calcium
| Chromium
I Cobalt
I Copper
I Iron
|Lead
| Magnesium
| Manganese
| Mercury
I Nickel
| Potassium
| Selenium
| Silver
| Sodium
| Thallium
| Vanadium
| Zinc
| Cyanide
1
Control
Limit
%R

























Spiked Sample
Result (SSR)

























C

























Sample
Result (SR)

























C

























Spike
Added (SA)

























%R

























Q

























M

























Comments:
                              FORM V (PART 1)  - IN
                                              3/90

-------
                                U.S.  EPA - CLP
                                       5B
                       POST DIGEST SPIKE SAMPLE  RECOVERY
                                     EPA SAMPLE NO.
                                                             I
Lab Name:

Lab Code:
Matrix  (soil/water):
               Contract:
Case No.:
SAS No.:
SDG No.:
                              Level  (low/med):
                          Concentration Units: ug/L
1
1
1
I Analyte
1
| Aluminum
| Antimony
I Arsenic
I Barium
| Beryllium
| Cadmium
I Calcium
I Chromium
I Cobalt
| Copper
| Iron
j Lead
I Magnesium
| Manganese
I Mercury
I Nickel
I Potassium
I Selenium
I Silver
1 Sodium
1 Thallium
I Vanadium
I Zinc
I Cyanide
1
Control
Limit
%R

























Spiked Sample
Result (SSR)

























                                   C|
                 Sample
              Result  (SR)  C|
Spike
Added (SA)

























%R

























Q
























_l
                                M
Comments:
                              FORM V (PART 2) - IN
                                              3/90

-------
                               U.S. EPA - CLP
                                  DUPLICATES
                                                            EPA  SAMPLE  NO,
                                                            r
Lab Name:



Lab Code:
Contract:
                       Case No.:
    SAS No.:
                      SDG No.:
Matrix (soil/water):



% Solids for Sample:
               Level (low/med):



        % Solids for Duplicate:
            Concentration Units (ug/L or mg/kg dry weight):
1
1
| Analyte
1
| Aluminum
| Antimony
j Arsenic
| Barium
| Beryllium
| Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
| Iron
|Lead
| Magnesium
| Manganese
| Mercury
| Nickel
| Potassium
| Selenium
(silver
| Sodium
| Thallium
j Vanadium
| Zinc
| Cyanide
1
Control
Limit

























Sample (S)

























C



















































Duplicate (D)

























C

























1
1
RPD


*






















Q

























1
M

























                                 FORM VI -  IN
                              3/90

-------
Lab Name:

Lab Code:
                                U.S. EPA - CLP
                           LABORATORY CONTROL SAMPLE
Contract:

SAS No.:
Case No.:
SDG No.
Solid LCS Source:

Aqueous LCS Source:
1
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
j Cadmium
| Calcium
| Chromium
I Cobalt
| Copper
llron
1 Lead 	
| Magnesium
j Manganese
j Mercury
| Nickel
| Potassium
j Selenium
| Silver
\ Sodium
| Thallium_
j Vanadium_
| Zinc
| Cyanide
1
Aqueous (ug/L)
True Found %R











































































Sol
True Found


















































      Solid (mg/kg)
          C       Limits
                                                                          %R
                                   FORM VII - IN
                            3/90

-------
      U.S.  EPA - CLP
            8
STANDARD ADDITION RESULTS
Lab Name:
Lab Code
.
Case No
• »
Contract
SAS No. :
•
SDG
No. :
Concentration Units: ug/L
EPA
Sample
No.

































An




~~















~~

—







0 ADD
ABS























'









1 ADD
CON ABS


































































2 ADD
CON ABS































































	

3 ADD
CON ABS
































































1
Final
Cone.

-




























1
!
r






























1


Q

-















"







—





       FORM VIII - IN
                                            3/90

-------
Lab Name:



Lab Code:
                               U.S.  EPA - CLP
                              ICP SERIAL DILUTIONS
               Contract:
                                                             EPA SAMPLE NO.
Case No.:
SAS No.:
SDG No.:
Matrix (soil/water):
                              Level  (low/med):
                          Concentration Units: ug/L
1
1
I Analyte
1
| Aluminum
| Antimony
{Arsenic
| Barium
1 Beryllium
| Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
I Iron
[Lead
| Magnesium
| Manganese
| Mercury
| Nickel
| Potassium
| Selenium
| Silver
| Sodium
| Thallium
| Vanadium
| Zinc
1
Initial Sample
Result (I)
























C

















































Serial
Dilution
Result (S)
























C

















































%
Differ-
ence
























1
1
1
1
1
I
1
1

i

















Q
























M






















1

                                  FORM IX - IN
                                              3/90

-------
Lab Name:

Lab Code:
ICP ID Number:

Flame AA ID Number:

Furnace AA ID Number:
                               U.S. EPA - CLP
                                     10
                   INSTRUMENT DETECTION LIMITS (QUARTERLY)
Contract:

SAS No.:

Date:
Case No.
                                      SDG No.
Comments:
1
1
I
| Analyte
1
| Aluminum
| Antimony
(Arsenic
| Barium
| Beryllium
| Cadmium
j Calcium
| Chromium
| Cobalt
| Copper
| Iron
|Lead
| Magnesium
| Manganese
| Mercury
j Nickel
| Potassium
| Selenium
| Silver
| Sodium
| Thallium
! Vanadium
| Zinc
Wave-
length
(nm)























Back-
ground























CRDL
(ug/L)
200
60
10
200
5
5
5000
10
50
25
100
3
5000
15
0.2
40
5000
5
10
5000
10
50
20
IDL
(ug/L)























                                                          M
                                  FORM X - IN
                           3/90

-------
                                U.S.  EPA - CLP

                                     11A
                ICP  INTERELEMENT CORRECTION FACTORS (ANNUALLY)
Lab Name:

Lab Code:
Case No.
ICP ID Number:
Contract:

SAS No.:

Date:
SDG No.
Analyte
Aluminum
Antimony-
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnes ium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc

Wave-
length
(nm)

















































Ii
Al
























rterelement
Ca
























Correction
Fe
























Factors foi
Mg

-






















<• *
























Comments:
                              FORM XI (PART 1)  - IN
                                              3/90

-------
                                U.S.  EPA - CLP

                                     11B
                ICP  INTERELEMENT CORRECTION FACTORS (ANNUALLY)
Lab Name:

Lab Code:
Case No.
ICP ID Number:
Contract:

SAS No.:

Date:
                                       SDG No.:
1
!
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic 	
| Barium
| Beryllium
| Cadmium 	
j Calcium
[ Chromium
I Cobalt
| Copper
| Iron
(Lead
[Magnesium
| Manganese
| Mercury
| Nickel
| Potassium
| S- lenium
[Silver
| Sodium
| Thallium
| Vanadium
| Zinc
1
Wave-
length
(nm)
























Interelement Correction Factors for:









































































-














































Comments:
                             FORM XI  (PART  2)  -  IN
                                              3/90

-------
                               U.S. EPA  -  CLP
                                     12
                        ICP LINEAR RANGES (QUARTERLY)
Lab Name:

Lab Code:
ICP ID Number:
Case No.:
Contract:

SAS No.:

Date:
SDG No.
1
1
1
| Analyte
1
| Aluminum
| Antimony
| Arsenic
| Barium
| Beryllium
1 Cadmium
| Calcium
| Chromium
| Cobalt
| Copper
llron
[Lead
(Magnesium
| Manganese
| Mercury
| Nickel
| Potassium
| Selenium
| Silver
| Sodium
| Thallium
| Vanadium
| Zinc
1
Integ.
Time
(Sec.)
























Concentration
(ug/L)
























M
























Comments:
                                  FORM  XII  - IN
                                              3/90

-------
                                  U.S.  EPA - CLP
                                        13
                                  PREPARATION LOG
Lab Name:

Lab Code:

Method:
Case No.:
Contract:

 SAS No.:
                                       SDG No.:
                EPA
               Sample
                No.
       Preparation
           Date
      Weight
      (gram)
Volume
 (mL)
                                   FORM XIII - IN
                                                       3/90

-------
   U.S.  EPA - CLP




           14




   ANALYSIS RUN LOG
Lab Name:
Lab Code
•
*
Case No.
i

Instrument ID Number:
Start Date:

EPA
Sample
No.


































D/F

































Time


































% R


































Contract
SAS No. :
Method:
End Date
»


i

SDG No. :




Analytes
A
L

































s
B

































|A
s

































IB
|A

































B
E

































1C
D

































1C
A
































c
R

































1C
o

































c
u

































F
E

































P
B

































M
G
































|M
N

































H
G































|N
I

































K

.































IS
E
























,







A
G

































N T V
A L|




!




















































































z c
N N










1












































FORM XIV - IN
                                            3/90

-------
    U.S.  EPA - CLP




           14




    ANALYSIS RUN LOG
Lab Name
: Contract:
Lab Code:
Case No . :


Instrument ID Number:
Start Date :

EPA
Sample
No.

















D/F


















1













1
Time

	































% R


































SAS No.:
SDG No. :
Method: 	
End Date:





Analytes
A S
L|B

































































|A
S































IB
IA
































1
B
IE

































1C
D

































1C
|A



























.





C
IR

































1C
10

































1C
|U

































IF
IE

































P
B































|M
G

































|M
N

































|H
IG































N
I

































K

































IS
E

































|A
IG

































IN
IA

































IT
IL

































IV
1

































12
IN































c
N

































FORM XIV - IN
                                           3/90

-------
SAMPLE LOG-IN SHEET
*,
Lab Name: P«RC of
Received BY (Print Nome): I ng-in n««r-
Reccived By (Signature): 	

Case Number:
Sample Delivery
Group No.:
SAS Number:

REMARKS:
1. Custody Seal(s) Present/Absent*
Intact/Broken
2 Custody Seal Notr

3. Caaim-of-Custody Present/Absent*

4. TruCfic Reports or Present/Absent'
Piciing Lift
: 5. AiibOl Aiibill/Sucter
i Prcsent/Abicol*
•: 6. AiifcOI No-

I 7. Simple T«gs Presea(/Absait*
: S«apIcT«c Listed/Not Listod
=. Numbers oa Ch«in-of-
< Custody
7 g. S^coplr Condiuon: Intact/Broken*/
Leaking
9. Does iafocmaboa oa
icpotte. «ad cample
• 10. DaleRecetvolatLab:

1 1. Tune Received:

Sample Transfer

Arcmft
By: 	 	
Oo:

EPA
SAMPLE
#


—






















CORRESPONDING
SAMPLE
TAG
#

























ASSIGNED
LAB
ff


























REMARKS:
CONDITION
OF SAMPLE
SHIPMENT. ETC.





-



















* CocUAd SMO wd >cuch irxxxxi of rcsoJoxioa
Rcvic-wed By:  _
D«lc:
Logbook  No.:  	
Logbook  P«gc No:
                                                      TORMOC-J
                                                                                                           3/90

-------
                               FULL INORGANICS
                           COMPLETE SDG FILE (CSF)
                               INVENTORY SHEET

Lab Name: 	           City/State:

Case No.        SDG No.        SDG Nos. to Follow: 	
SAS No.         Contract No.           SOW No.
All documents delivered in the Complete SDG File must be original documents
where possible.  (Reference Exhibit B, Section II D and Secti'on III V)
                                                  Page  Nos.    (Please Check:)
                                                 From    To      Lab  Region
1.   Inventory Sheet (DC-2) (Do not number)
2.   Cover Page
3.   Inorganic Analysis
     Data Sheet (Form I-IN/
4.   Initial & Continuing Calibration
     Verification  (Form IIA-IN)
5.   CRDL Standards For AA and ICP
      (Form IIB-IN)
6.   Blanks  (Form III-IN)
7.   ICP Interference Check
     Sample  (Form IV-IN)
8.   Spike Sample Recovery  (Form VA-IN)
9.   Post Digest Spike
     Sample Recovery (Form VB-IN)
10.  Duplicates  (Form VI-IN)
11.  Laboratory Control Sample
      (Form VII-IN)
12.  Standard Addition Results
     (Form VIII-IN)
13.  ICP Serial Dilutions  (Form IX-IN)
14.  Instrument Detection Limits
     (Form X-IN)
15.  ICP Interelement Correction Factors
     (Form XIA-IN)
16.  ICP Interelement Correction Factors
     (Form XIB-IN)
17.  ICP Linear Ranges (Form XII-IN)
18.  Preparation Log (Form XIII-IN)
19.  Analysis Run Log (Form XIV-IN)
20.  ICP Raw Data
21.  Furnace AA Raw Data
22.  Mercury Raw Data
                                  Form DC-2                              3/90

-------
                                                  Page Nos.
                                                 From    To
                                                          (Please  Check:)
                                                            Lab   Region
23.
24.
25.
26.
27.
28.
29.
30.
31.
Cyanide Raw Data
Preparation Logs Raw Data
Percent Solids Determination Log
Traffic Report
EPA Shipping/Receiving Documents
   Airbill  (No. of Shipments 	)
   Chain-of-Custody Records
   Sample Tags
   Sample Log-In Sheet (Lab & DC1)
   SDG Cover Sheet
Misc. Shipping/Receiving Records
(list all individual records)
   Telephone Logs
Internal Lab Sample Transfer Records &
Tracking Sheets (describe or list)
Internal Original Sample Prep & Analysis Records
(describe or list)
   Prep Records 	           	
   Analysis Records 	         	
   Description 	          	
Other Records (describe or list)
   Telephone Communication Log
32.  Comments:
Completed by (CLP Lab):
          (Signature)

Audited by (EPA):
                                 (Print Name & Title)
(Date)
           (Signature)
                                 (Print Name & Title)
(Date)
                            Form DC-2 (continued)
                                                                    3/90

-------
          EXHIBIT C
INORGANIC TARGET ANALYTE LIST
                                                ILM01.0

-------
                        INORGANIC TARGET ANALYTE LIST (TAL)
                                            Contract Required
                                              Detection Limit ^ '2'
          Analyte                                 (ug/L)
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide
200
60
10
200
5
5
5000
10
50
25
100
3
5000
15
0.2
40
5000
5
10
5000
10
50
20
10
(1)   Subject to the restrictions  specified in  the first page  of  Part  G,
     Section IV of Exhibit D  (Alternate Methods  - Catastrophic Failure)  any
     analytical method specified  in SOW Exhibit  D may be utilized  as  long as
     the documented instrument  or method detection  limits meet the Contract
     Required Detection Limit (CRDL)  requirements.  Higher  detection  limits
     may only be used in the  following circumstance:

               If  the  sample  concentration exceeds five times the  detection
               limit of  the instrument or method in use,  the value may be
               reported  even  though the instrument or method detection limit
               may not equal  the  Contract Required Detection Limit.  This  is
               illustrated in the  example below:

               For lead:

               Method  in use  =  ICP
               Instrument  Detection Limit (IDL)  =.40
               Sample  concentration = 220
               Contract  Required  Detection Limit (CRDL) = 3
                                   C-l                                 ILM01.0

-------
              The value of 220 may be reported even though instrument
              detection limit is greater than CRDL. The instrument or
              method detection limit must be documented as described in
              Exhibit E.

(2)   The  CRDL are  the instrument detection limits obtained in pure water
     that must be  met using the procedure in Exhibit E.  The detection
     limits for samples may be considerably higher  depending on the sample
     matrix.
                                   C-2                                ILM01.0

-------
                                 EXHIBIT D
                             ANALYTICAL METHODS
                                                                  Page No.
SECTION I   - INTRODUCTION	   D-l
     Figure 1-Inorganic Methods Flow Chart  	   D-3
SECTION II  - SAMPLE PRESERVATION AND HOLDING TIMES	   D-4
     Part A - Sample Preservation   	- D-4
     Part B - Holding Times	   D-4
SECTION III - SAMPLE PREPARATION	   D-5
     Part A - Water Sample Preparation	   D-5
     Part B - Soil/Sediment Sample Preparation  	   D-6
     Part C - Microwave Digestion Method	   D-9
     Part D - Mercury and Cyanide Preparation	     D-16
SECTION IV  - SAMPLE ANALYSIS	   D-17
     Part A - Inductively Coupled Plasma-Atomic,
              Emission Spectrometric Method .    	   D-18
     Part B - Atomic Absorption Methods,  Furnace Technique  .  .  .   D-32
     Part C - Atomic Absorption Methods,  Flame Technique  ....   D-45
     Part D - Cold Vapor Methods for Mercury Analysis	   D-50
     Part E - Methods for Total Cyanide Analysis  	   D-69
     Part F - Percent Solids Determination Procedure  	   D-98
     Part G - Alternate Methods (Catastrophic  ICP Failure)   .  .'  .   D-99
                                                                      ILM01.0

-------
                                  SECTION I

                                 INTRODUCTION
 Inorganic Methods Flow Chart:   Figure I outlines the general  analytical
 scheme the Contractor will follow in performing analyses  under this
 contract.

 Permitted Methods:   Subject to the restrictions specified in  Section IV,
 Part G -  Alternate  Methods (Catastrophic ICP Failure),  any analytical
 method specified in Exhibit D  may be used as long as the  documented
 instrument or method detection limits meet the Contract Required Detection
 Limits (Exhibit C).  Analytical methods with higher detection  limits may be
 used only if the sample concentration exceeds five times  the  documented
 detection limit of  the instrument or method.

 Initial Run Undiluted:  All samples must initially be run undiluted (i.e.,
 final product of the sample preparation procedure).   When an  analyte
 concentration exceeds the calibrated or linear range (as  appropriate),  re-
 analysis  for that analyte(s) is required after appropriate dilution.  The
 Contractor must use the least  dilution necessary to bring the analyte(s)
 within the valid analytical range (but not below the CRDL) and repo'rt the
 highest valid value for each analyte as measured from the undiluted and
 diluted analyses.   Unless the  Contractor can submit proof that dilution was
 required to obtain  valid results, both diluted and undiluted  sample
 measurements must be contained in the raw data.  All sample dilutions shall
 be  made with deionized water appropriately acidified to maintain constant
 acid strength.

 Quality Assurance/Quality Control Measurements:  The Contractor is  reminded
 and cautioned that  Exhibit D is a compendium of required  and/or permitted
 analytical methods  to be used in the performance of analyses  under  this
 contract.   The quality assurance/quality control procedures or measurements
 to  be performed in  association with these methods or analyses are specified
 in  Exhibit E.  In the event references to quality assurance measurements in
 any of the methods  appear to be in conflict with or to be less stringent
 than the requirements of Exhibit E, the requirements of Exhibit E will
 prevail.

 Raw Data Requirements:  The Contractor is reminded and cautioned that the
 collection and provision of raw data may or may not be referred to  within
 the individual methods of Exhibit D or the Quality Assurance  Protocol of
-Exhibit E.  The Raw Data Deliverables requirements are specified in Exhibit
 B,  Section II.D.2.d.  Raw data collected and provided in  association with
 the performance of analyses under this contract shall conform to the
 appropriate provisions of Exhibit B.

 Glassware Cleaning:   Lab glassware to be used in metals analysis must be
 acid cleaned according to EPA's manual "Methods for Chemical  Analysis of
 Water and Wastes" or an equivalent procedure.

 Standard Stock Solutions:  Stock solutions to be used for preparing
 instrument or method calibration standards may be purchased or prepared as
 described in the individual methods of Exhibit D.  All other  solutions to
 be  used for Quality Assurance/Quality Control measurements shall conform  to
 the specific requirements of Exhibit E.

                                     D-l                                 ILM01.0

-------
Aqueous  Sample pH Measurement:  Before sample preparation is initiated on
an aqueous sample received in shipment, the Contractor must check the pH of
the sample and note in a preparation log if the pH is <2 for a metals
sample or if the pH is >12 for a cyanide sample.   The Contractor shall not
take any pH adjustment action if the sample has not been properly
preserved.

Sample Mixing:  Unless instructed otherwise by the EPA Administrative
Project  Officer or Technical Project Officer,  all samples shall be mixed
thoroughly prior to aliquoting for digestion.   No specific procedure is
provided herein for homogenization of soil/sediment samples; however, an
effort should be made to obtain a representative  aliquot.

Background Corrections:   Background corrections are required for Flame AA
measurements below 350 nm and for all Furnace  AA measurements.   For ICP
background correction requirements,  see Exhibit D Section IV,  Part A,
paragraph 2.1.

Replicate Injections/Exposures:   Each furnace  analysis requires a minimum
of two injection (burns),  except for full method  of Standard Addition
(MSA).   All ICP measurements  shall require  a minimum of two -aplicate
exposures.   Appropriate  hard copy raw data  for each exposure/injection
shall  be included in the data package in accordance with Exhibit B,* Section
II,  Part D,  paragraph 2.d.   The  average of  each set of exposures/injections
shall  be used for standardization,  sample analysis,  and reporting as
specified in Exhibit D.
                                   D-2                                ILM01.0

-------
            Figure 1

  INORGANICS METHODS  FLOW CHART
    I                  I
    |  Field Sample    |
           .1.
  Traffic Report or SMO
   Specifies Parameters
Water
Matrix
Soil/Sediment
    Matrix
                  I
Cyanide | |Acid Digestion) |Acid Digestion| |% Solids | | Cyanide
Analysis | | for Metals | | for Metals j |Determin-| | Analysis
in Water | | Analysis | | Analysis in | | ation | | in Soil/
| | in Water | (Soil/Sediment | | | (Sediment
! 1 1
1 1 1
1 1 1
| [Metal Anal.j (Metals Anal.J
j | ICP/AAS | | 1CP/AAS |
1 1 1
1 1 1
I 1 1
! 1 1
I | Data Reports |
1
1
1
1
1
1
!
1
!
             D-3
                                      ILM01.0

-------
                            SECTION II

               SAMPLE PRESERVATION AND HOLDING TIMES
Sample Preservation
     Water Sample Preservation
     Measurement
      Parameter
     Metals(3)
     Cyanide,  total
       and amenable
       to chlorination
     FOOTNOTES:
Container' ^

   P.G
   P,G
 Preservative' '
HN03 to pH <2

0.6g ascorbic acid(4)
NaOH to pH >12
Cool, maintain at 4°C(±2°C)
 until analysis
     (1)   Polyethylene (P)  or glass (G).
     (2)   Sample  preservation is  performed by  the  sampler  immediately
          upon  sample collection.
     (3)   Samples are filtered immediately on-site by  the  sampler
          before  adding preservative for  dissolved metals.
     (4)   Only  used in the  presence of residual  chlorine.

     Soil/Sediment Sample Preservation

     The  preservation required for soil/sediment samples is maintenance
     at 4°C (± 2°) until analysis.
Holding Times for Water and Soil/Sediment Samples

Following are the maximum sample holding times allowable under this
contract.  To be compliant with this contract, the Contractor must
analyze samples within these times even if these times are less than
the maximum data submission times allowed in this contract.
       Analvte


     Mercury
     Metals (other than mercury)
     Cyanide
             No. of Days Following
                 Sample Receipt
                 by Contractor

                      26 days
                     180 days
                      12 days
                               D-4
                                             ILM01.0

-------
                                 SECTION III

                             SAMPLE PREPARATION


A.   WATER SAMPLE PREPARATION

     1.    Acid Digestion Procedure for Furnace Atomic Absorption Analysis

          Shake sample and transfer 100 mL of well-mixed sample  to a 250-mL
          beaker,  add 1 mL of (1+1) HN03 and 2 mL 30% H202  to  the sample.
          Cover with watch glass or similar cover and heat  on  a  steam bath
          or hot plate for 2 hours at 95°C or until sample  volume is reduced
          to between 25 and 50  mL, making certain sample does  not boil.
          Cool sample and filter to remove insoluble material.   (NOTE:  In
          place of filtering,  the sample,  after dilution and mixing, may be
          centrifuged or allowed to settle by gravity overnight  to remove
          insoluble material.)  Adjust sample  volume to 100  mL  with deionized
          distilled water.   The sample is  now ready for analysis.

          Concentrations so determined shall  be reported as "total".

          If Sb is to be determined by furnace AA,  use the  digestate
          prepared for ICP/flame AA analysis.

     2.    Acid Digestion Procedure for ICP and Flame AA Analyses

          Shake sample and transfer 100 mL of well-mixed sample  to a 250-mL
          beaker,  add 2 mL of (l-i-l) HNO-j and 10 mL of (1+1) HC1  to the
          sample.   Cover with watch glass  or  similar cover  and heat on a
          steam bath or hot plate for 2 hours at 95°C or until sample volume
          is reduced to between 25 and 50  mL,  making certain sample does not
          boil.  Cool sample and filter to remove insoluble material.
          (NOTE:  In place of filtering,  the sample,  after dilucion and
          mixing,  may be centrifuged or allowed to  settle by gravity
          overnight to remove  insoluble material.)  Adjust sample  volume  to
          100 mL with deionized distilled  water.   The sample is  now ready
          for analysis.

          Concentrations so determined shall  be reported as "total".
                                    D-5                                 ILMOL.O

-------
B.    SOIL/SEDIMENT  SAMPLE  PREPARATION
     1.   Acid Digestion Procedure for ICP, Flame AA and Furnace AA Analyses

         a.   Scope and Application

              This method is an acid digestion procedure used to prepare
              sediments,  sludges,  and soil samples for analysis by flame or
              furnace atomic absorption spectroscopy (AAS) or by
              inductively coupled plasma spectroscopy (ICP)'.   Samples
              prepared by this method may be analyzed by AAS or ICP for the
              following metals:
                Aluminum
                Antimony
                Arsenic
                Barium
                Beryllium
                Cadmium
                Calcium
              Summary of Method
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
              A representative  1  g (wet weight)  sample is digested in
              Nitric acid and hydrogen peroxide.   The digestate is then
              refluxed with either Nitric acid or Hydrochloric acid.
              Hydrochloric acid is used as the final reflux acid for the
              furnace AA analysis of Sb,  the Flame AA or ICP analysis of
              Al,  Sb, Ba,  Be, Ca,  Cd,  Cr, Co,  Cu, Fe, Pb, Mg,  Mn,  Ni,  K,
              Ag,  Na, Tl,  V and Zn.   Nitric acid is employed as the final
              reflux acid for the Furnace AA analysis of As,  Be,  Cd,  Cr,
              Co,  Cu, Fe,  Pb, Mn,  Ni,  Se, Ag,  Tl, V,  and Zn.   A separate
              sample shall be dried for a percent solids determination
              (Section IV,Part  F).

              Apparatus and Materials

              (1)   250  mL beaker  or  other appropriate vessel.

              (2)   Watch glasses

              (3)   Thermometer  that  covers range  of 0° to 200°C

              (4)   Whatman No.  42  filter  paper or equivalent

              Reagents

              (1)   ASTM Type  II water  (ASTM D1L93):   Water must be
                   monitored.

              (2)   Concentrated nitric  acid (sp.  gr.  1.41)

              (3)   Concentrated Hydrochloric Acid (sp.  gr. 1.19)

                                   D-6                                 ILM01.0

-------
     (4)  Hydrogen Peroxide  (30%)

e.   Sample  Preservation  and Handling

     Soil/sediment  (nonaqueous)  samples muse be  refrigerated ac
     4°C  (±2°)  from  receipt  until analysis.

f.   Procedure

     (1)  Mix the sample  thoroughly to achieve homogeneity.   For
         each digestion  procedure, weigh  (to the nearest O.Olgms)
         a 1.0 to 1.5 gm portion of sample  and  transfer to  a
         beaker,

     (2)  Add 10 mL of 1:1 nitric acid (HNO-j) , mix the slurry, and
         cover with  a watch glass.  Heat the sample to 95°C and
         reflux for 10 minutes without boiling.   Allow the  sample
         to cool, add 5 mL of concentrated  HNO-j, replace the
         watch glass,  and reflux for 30 minutes.  Do not allow
         the volume to be reduced to less than  5 mL while
         maintaining a covering of solution over the bottom of
         the beaker.

     (3)  After the second reflux step  has been completed and the
         sample has cooled,  add 2 mL of Type II  water and 3 mL of
         30% hydrogen peroxide (1^02).   Return the beaker to the
         hot plate for warming to start the  peroxide reaction.
         Care must be taken to ensure  that  losses do not  occur
         due to excessively vigorous effervescence.   Heat until
         effervescence subsides, and cool the beaker.

     (4)  Continue to add 30% #2®2 ^n ^  mL aliquots with warming
         until  the  effervescence is minimal or  until the general
         sample appearance is   unchanged.    (NOTE:  Do not add
         more than a total of 10 mL 30% t^C^. )

     (5a)  If the sample is being prepared for the furnace AA
         analysis of Sb,  the flame  AA or  ICP analysis of Al, Sb,
         Ba,  Be,  Ca,  Cd,  Cr, Co,  Cu,  Fe,  Pb,  Mg, Mn, Ni,  K, Ag,
         Na,  Tl,  V,  and  Zn,  add 5 mL of 1:1 HC1  and  10 mL of Type
         II  water,  return the  covered beaker to  the  hot plate,
         and heat for  an additional 10  minutes.   After cooling,
         filter through  Whatman No.  42  filter paper  (or
         equivalent)  and  dilute to  100  mL with Type  II  water.
         NOTE:   In place  of  filtering,  the sample (after dilution
         and mixing)  may  be  centrifuged or allowed to settle by
         gravity  overnight to  remove insoluble material.The
         diluted  sample  has  an approximate acid  concentration of
         2.5%  (v/v)  HC1  and  5% (v/v) HN03.   Dilute  the  digestate
         1:1  (200 mL final volume)  with acidified water Co
         maintain constant acid strength.   The sample is  now
         ready  for  analysis.
                         D-7                                ILM01.0

-------
     (5b) If the sample is being prepared for  the  furnace analysis
          of As,  Be,  Cd,  Cr, Co,  Cu,  Fe,  Pb, Mn, Ni,  Se,  Ag,  Tl,
          V,  and Zn,  continue heating the acid-peroxide digestate
          until  the volume has been  reduced  to approximately  2 raL,
          add 10 raL of Type II water,  and warm the  mixture.  After
          cooling, filter through Whatman No. 42 filter paper (or
          equivalent)  and dilute the sample  to 100  mL with Type II
          water  (or centrifuge the sample).  NOTE:   In place  of
          filtering,  the  sample  (after dilution and mixing) may be
          centrifuged or  allowed Co  settle by gravity overnight to
          remove insoluble material.   The diluted  digestate
          solution contains approximately 2% (v/v)  HNO-,.   Dilute
          the digestate 1:1 (200 mL  final volume)  with acidified
          water  to maintain constant acid strength.   For
          analysis, withdraw aliquots  of  appropriate volume,  and
          add any required reagent or matrix modifier.  The sample
          is now ready for analysis.

g.   Calculations

     (1)   A separate  determination of percent solids must be
          performed  (Section IV,  Part F).

     (2)   The concentrations determined in the digest are to  be
          reported on the basis  of the dry weight of the  sample.

              Concentration (dry  wt.) (mg/kg) =  C  x V
                                                 W  x S

              Where,
                 C =   Concentration   (mg/L)
                 V =   Final  volume in  liters after  sample
                      preparation
                 W =   Weight in  kg of  wet sample
                 S =   % Solids/100

h.   Bibliography

     Modification (by committee) of  Method 3050,  SW-846,  2nd ed.,
     Test Methods for Evaluating Solid Waste.  EPA  Office  of Solid
     Waste and  Emergency Response,  July 1982.
                          D-8                                 ILM01.0

-------
 C.   TOTAL METALS SAMPLE PREPARATION USING MICROWAVE DIGESTION

 1.     SCOPE AND  APPLICATION

       This method is an acid digestion procedure using microwave energy  to
       prepare water and soil samples for  analysis by GFAA,  ICP, or Flame AA
       for the following metals:
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium '
Silver
Sodium
Thallium
Vanadium
Zinc
2.    SUMMARY OF METHOD

      a.      Water  Sample  Preparation

              A representative  45  mL water  sample  is  digested  in  5 raL  of
              concentrated  Nitric  acid in a Teflon PFA vessel for 20
              minutes using microwave heating.  The digestate  is  then
              filtered  to remove insoluble  material.   The  sample  may be
              centrifuged or allowed to settle by  gravity  overnight to
              remove insoluble  material.

      b.      Soil Sample Preparation

              A representative  0.5 g (wet weight)  sample is digested in 10
              mL of concentrated Nitric acid in a  Teflon   PFA vessel for 10
              minutes using microwave heating.  The digestate  is  then
              filtered  to remove insoluble  material.   The  sample  may be
              centrifuged or allowed to settle by  gravity  overnight to
              remove insoluble  material.

3.    APPARATUS  AND MATERIALS

      a.      Commercial kitchen or  home-use microwave ovens shall not be
              used for  the  digestion of samples under  this contract.  The
              oven cavity must  be corrosion resistant  and well ventilated.
              All electronics must be  protected against corrosion for  safe
              operation.

      b.      Microwave oven with programmable power settings up  to at
              least 600 Watts.
                                            n
      c.      The system must use PFA Teflon  digestion vessels (120 mL
              capacity)  capable of withstanding pressures of up to 110 zlO
              psi (7.5 ±0.7  atm).   These  vessels are capable of controlled
              pressure relief at pressures  exceeding 110 psi.
                                   D-9                                 ILM01.0

-------
      d.      A rotating turntable must be used to ensure homogeneous
              distribution of microwave radiation within the oven.  The
              speed of the turntable must be a minimum of 3 rpm.
                                                          T^
      e.      Polymeric volumetric ware in plastic (Teflon  or
              polyethylene) 50 mL or 100 mL capacity.

      f.      Whatman No. 41 filter paper (or equivalent).

      g.      Disposable polypropylene filter funnel.

      h.      Analytical balance, 300 g capacity, and minimum ±0.01 g.

      i.      Polyethylene bottles, 125 mL, with caps.

4.    REAGENTS

      a.      ASTM Type II water (ASTM D1193):  water must be monitored.

      b.      Sub-boiled, concentrated Nitric Acid (sp. gr. 1.41)..

      c.      Concentrated Hydrochloric Acid (sp. gr. 1.19).

5.    MICROWAVE CALIBRATION PROCEDURE

      a.      The calibration procedure is a critical step prior  to the use
              of any microwave unit.  The microwave unit must be  calibrated
              every six months.  The calibration data for each  calibration
              must be available for review during on-site audits.  In  order
              that absolute power settings may be interchanged  from one
              microwave unit to another, the actual delivered power must  be
              determined.

              Calibration of a laboratory microwave unit depends  on the
              type of electronic system used by the manufacturer.  If  the
              unit has a precise and accurate linear relationship between
              the output power and the scale used in controlling  the
              microwave unit, then the calibration can be a two-point
              calibration at maximum and 40% power.  If the unit  is not
              accurate or precise for some portion of the controlling
              scale,  then a multiple-point calibration is necessary.   If
              the unit power calibration needs a multiple point
              calibration, then  the point where  linearity begins  must  be
              identified.  For example:  a calibration at  100,  99, 98, 97,
              95, 90, 80, 70, 60, 50 and 40% power settings can be applied
              and the data plotted.  The non-linear portion of  the
              calibration curve  can be excluded  or restricted  in  use.  Each
              percent  is equivalent  to approximately  5.5  - 6 watts and
              becomes  the  smallest unit of power that can  be controlled.
               If  20  - 40 watts are contained from 99-100%, that portion  of
               the microwave calibration is not controllable by  3-7 times
               that  of the  linear portion of  the  control scale  and will
               prevent duplication  of precise power conditions  specified  in
               that  portion of  the  power scale.

                                   D-10                                ILM01.0

-------
         The power available for heating  is evaluated so  that the
         absolute power setting (watts) may be compared  from one
         microwave to  another.  This is accomplished by measuring the
         temperature rise  in 1 Kg of water exposed to microwave
         radiation for a fixed period of  time.  The water is placed in
         a  teflon  beaker  (or a beaker that is made of some  other
         material that does not adsorb microwave  energy)  and stirred
         before measuring  the temperature.  Glass beakers adsorb
         microwave energy  and may not be  used.  The initial
         temperature of the water must be between 19 and  25  "C.   The
         beaker is circulated continuously through the field for  at
         least two (2) minutes at full power.  The beaker is removed
         from the microwave, the water is stirred vigorously,  and the
         final temperature recorded.  The final reading  is the maximum
         temperature reading after each energy exposure.   These
         measurements  must be accurate to ± 0.1 "C and made  within 30
         seconds of the end of heating.   If more  measurements are
         needed, do not use the same water until  it has cooled down to
         room temperature.  Otherwise, use a  fresh water  sample.

         The absorbed  power is determined by  the  following formula:

                   P _ (tO rCo") Cm) (DT)
Where:

P - The apparent power absorbed by the sample in watts (joules per
        second),

K = The conversion factor for thermochemical calories per second to
        watts  (-4.184),

C  = The heat capacity, thermal capacity, or specific heat (cal. g"
        1.°C'1) of water  (-1.0),

m — The mass of the sample in grams (g),

DT = the final temperature minus the initial tempereture (°C), and

t = the time in seconds (s)

Using 2 minutes and 1 Kg of distilled water, the calibration equation
simplifies to:

P = (DT) (34.87).

The microwave user can now relate power in watts to the percent power
setting of the microwave
                             D-ll                                ILK01.0

-------
6.    CLEANING PROCEDURE

a.    The  initial cleaning of  the  PFA vessels:

(1)   Prior to first use  - new vessels must be annealed before  they are
      used.  A pretreatment/cleaning procedure must be followed.   This
      procedure calls for heating  the vessels for  96 hours  at 200°C.  The
      vessels must be disassembled during annealing and the sealing
      surfaces (the top of the vessel or its rim)  must not  be used to
      support the vessel during annealing.

(2)   Rinse in ASTM Type I water.

(3)   Immerse in 1:1 HC1 for a minimum of 3 hours  after the cleaning bath
      has  reached a temperature just below boiling.

(4)   Rinse in ASTM Type I water.

(5)   Immerse in 1:1 HNO-j for a minimum of 3 hours after the cleaning bath
      has  reached a temperature just below boiling.

(6)   The vessels are then rinsed with copious amounts of ASTM Type I water
      prior to use for any analyses under this contract.

b.    Cleaning procedure between sample digestions

(1)   Wash entire vessel in hot water using laboratory-grade nonphosphate
      detergent.

(2)   Rinse with 1:1 nitric acid.

(3)   Rinse three times with ASTM Type I water.   If contaminants are found
      in the preparation blank, it is mandatory that steps  a(2)  through
      a(6) be strictly adhered to.

7.    DIGESTION PROCEDURE

a.    Water Sample Digestion Procedure
                                                            P
(1)   A 45 mL aliquot of the sample are measured into Teflon  digestion
      vessels using volumetric glassware.

(2)   5  mL of high purity concentrated HNO-j is added to the digestion
      vessels.

(3)   The weight  of each vessel is recorded to 0.02 g.

(4)   The caps with the pressure release valves are placed  on the vessels
      hand tight  and then tightened,  using constant torque, to 12 ft./^bs.
      Place 5 sample vessels in the carousel,  evenly spaced around its
      periphery in the microwave unit.   Venting tubes connect each sample
      vessel with a collection vessel.   Each sample vessel  is attached to a
      clean, double-ported vessel to collect any sample expelled from the
      sample vessel in the event of over pressurization.   Assembly of the

                                   D-12                                ILM01.0

-------
vessels into the carousel may be done inside or outside the
microwave.

(5)     This procedure  is energy balanced for  five 45  mL  water
        samples  (each with 5 mL of acid) to produce  consistent
        conditions,  when fewer than 5 samples are digested,  the
        remaining vessels must be filled with  45 mL  of tap,  DI  or
        Type II  water and 5 mL of concentrated Nitric  acid.

        Newer microwave ovens may be capable of higher power settings
        which may allow a larger number of samples.  If the  analyst
        wishes to digest more than 5 samples at a time, the  analyst
        may use  different power settings as long as  they  result in
        the same time temperature conditions defined in the  power
        programming for this method.

        The initial temperature of the samples should  be  24  + 1°C.
        The preparation blank must have 45 mL of deionized water  and
        the same amount (5 mL) of acid that is added to the  samples.

        The microwave unit first-stage program must  be  set to give
        545 watts for 10 minutes and the second-stage  program to  give
        344 watts for 10 minutes.   This sequence brings the  samples
        to 160 ±4"C in ten minutes and permits a slow  rise to 165-170
        °C during the second 10 minutes.

(6)     Following the 20 minute program, the samples are  Left to  cool
        in the microwave unit for five  minutes,  with  the exhaust fan
        ON.  The samples and/or carousel may then be removed  from the
        microwave unit.   Before opening the vessels  let cool  until
        they are no longer hot to the touch.

(7)     After the sample vessel has cooled,  weigh the  sample vessel
        and compare to the initial weight as reported  in  the
        preparation log.  Any sample vessel exhibiting a < 0.5  g,  loss
        must have any excess sample from the associated collection
        vessel added to the original sample vessel before proceeding
        with the sample preparation.   Any sample vessel exhibiting a
        > 0.5 g loss must be identified in the preparation log  and
        the sample redigested.

(9)     Sample Filtration:

        The digested samples are shaken well to mix  in any condensate
        within the digestion vessel before being opened.  The
        digestates are then filtered into 50 mL glass volumetric
        flasks through ultra-clean filter paper and diluted to  50 mL
        (if necessary).   The samples are now ready for analysis.  The
        sample results must be corrected by a factor of 1.11 in order
        Co report final  concentration values based on an initial
        volume of 45 ml.  Concentrations so determined shall be
        reported as "total".
                             D-13                                 ILM01.0

-------
b.      Soil  Sample Digestion  Procedure

(1)     Add a representative 0.5 ±0.050 grams of sample  to  the
        TeflonR  PFA vessel.

(2)     Add 10 ±0.1 mL of concentrated nitric acid.   If  a vigorous
        reaction occurs, allow the reaction to stop before  capping
        the vessel.

(3)     Cap the  vessel, then tighten using constant torque  to 12
        ft/lbs,  according to the manufacturer's direction.

(4)     Connect  the sample vessel to the overflow vessel using
        TeflonR  PFA tubing.

(5)     Weigh the vessel assembly to the nearest O.Olg.

(6)     Place sample vessels in groups of 2 sample vessels  or 6
        sample vessels in the  carousel, evenly spaced around its
        periphery in the microwave unit.  If fewer than  the -
        recommended number of  samples are to be digested, i.e. 3
        samples  plus 1 blank then the remaining vessels  must be
        filled with 10 mL of water to achieve to the  full complement
        of vessels.

        Each  sample vessel must be attached to a clean,  double-ported
        vessel to collect any  sample expelled from the sample vessel
        in the event of over pressurization.  Assembly of the vessels
        into  the carousel may  be done inside or outside  the
        microwave.  Connect the overflow vessel to the center well  of
        the oven.

(7)     The preparation blank  must have 0.5 mL of deionized water and
        the same amount (10 mL) of acid that is added to the samples.
        The preparation blank  must later be diluted to 50 mL in the
        same  manner as the samples.

(8)     Irradiate the 2 sample vessel group at 344 watts for 10
        minutes, or the 6-sample vessel group at 574  watts  for 10
        minutes.

        This  program brings the samples to 175°C in 5.5  minutes, and
        remains  between 170-180°C for the balance of  the 10 minute
        irradiation period.  The pressure should peak at less than  6
        atm for  most samples.  The pressure may exceed these  limits
        in the case of high concentrations of carbonate  or  organic
        compounds.  In these cases,  the pressure will be limited by
        the relief pressure of the vessel to 7.5 ±0.7 atm.
                             D-14                                ILM01.0

-------
(9)     Allow  the  vessels  to  cool  for  a  minimum  of  five  minutes
        before removing  them  from  the  microwave  unit,  with exhaust
        fan ON.  Allow the vessels to  cool  to  room  temperature before
        opening.   The  vessels must be  carefully  uncapped and vented
        in a. fume  hood.

(10)    Weigh  each vessel  assembly.  If  the  weight  of  acid plus  the
        sample has decreased  by  more than 10%  from  the original
        weight, discard  the digests.   Determine  the reason for the
        loss.   Losses  typically  are attributed to use  of digestion
        time longer  than ten  minutes,  using  too  large  of a sample,  or
        having improper  heating  conditions.  Once the  source  of  the
        losses has been  corrected,  prepare a new set of  samples  for
        digestion.

(11)    Sample Filtration:

        Shake  the  sample well to mix in  any  condensate within the
        digestion  vessel before  being  opened.  Filter  the digestion
        vessel into  a  50 mL glass  volumetric flask  through ultra-
        clean  filter paper.   Rinse the sample  digestion  vessel,  cap,
        connecting tube, and  (if venting occurred)  the overflow
        vessel into  the  50 mL glass flask.   Dilute  to  50 mL.  The
        samples are  now  ready for  analysis.  Concentrations  so
        determined shall be reported as  "total".

(12)    Calculations:

        The concentrations determined  in the digest are  to be
        reported on  the  basis of the dry weight  of  the sample.
                  Concentration (dry wt.) (mg/Kg) =  C x V
                                                     W x S

                  Where
                    C =  Concentration  (mg/L)
                    V =  Final volume in liters after sample
                         preparation
                    U =  Weight in Kg of wet sample
                    S =  % Solids/100
                             D-15                                ILM01.0

-------
D.   MERCURY AND CYANIDE PREPARATION

      Refer Co each specific method in this  Exhibit for mercury and cyanide
      preparations.
                                   D-16                                ILM01.0

-------
                         SECTION  IV


                      SAMPLE ANALYSIS



                                                          .Page No.

PART A - INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION
         SPECTROMETRIC METHOD                                D-18

PART B - ATOMIC ABSORPTION METHODS,  FURNACE TECHNIQUE        D-32

PART C - ATOMIC ABSORPTION METHODS,  FLAME TECHNIQUE          D-45

PART D - COLD VAPOR METHODS FOR MERCURY ANALYSIS             D-50

PART E - METHODS FOR CYANIDE ANALYSIS                        D-69

PART F - PERCENT SOLIDS DETERMINATION PROCEDURE              D-98

PART G - ALTERNATE METHODS (CATASTROPHIC ICP FAILURE)        D-99
                         D-17                                ILM01.0

-------
   PART A -  INDUCTIVELY COUPLED  PLASMA-ATOMIC  EMISSION SPECTROMETRIC METHOD'*'


                             Method 200.7  CLP-M*
        INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION  SPECTROMETRIC METHOD
                FOR TRACE ELEMENT ANALYSIS OF WATER AND WASTES


1.    Scope and Application
                                                            •
1.1   Dissolved elements are determined in filtered and acidified samples.
      Appropriate steps must be  taken in all analyses to ensure that potential
      interferences are taken into account.  This  is especially true when
      dissolved solids exceed 1500 mg/L.  (See 5.)

1.2   Total elements are determined after appropriate digestion procedures are
      performed.  Since digestion techniques increase the dissolved -solids
      content of the samples, appropriate steps must be taken to correct for
      potential interfere-ce effects.  (See 5.)

1.3   Table 1 lists elements along with recommended wavelengths and'typical
      estimated instrumental detection limits using conventional pneumatic
      nebulization.  Actual working detected limits are sample dependent and
      as the sample matrix varies, these concentrations may also vary.  In
      time,  other elements may be added as  more information becomes available
      and as required.

1.4   Because of the differences between various makes and models of
      satisfactory instruments,  no detailed instrumental operating
      instructions can be provided.   Instead,  the  analyst is referred to the
      instructions provided by the manufacturer of the particular instrument.

2.    Summary of Method

2.1   The method describes a technique for  the simultaneous or sequential
      multielement determination of trace elements  in solution.  The basis of
      the method is the measurement of atomic emission by an optical
      spectroscopic technique.   Samples are nebulized and the aerosol that is
      produced is transported to the plasma torch where excitation occurs.
      Characteristic atomic-line emission spectra  are produced by a radio-
      frequency inductively coupled plasma (ICP).   The spectra are dispersed
      by a grating spectrometer  and the intensities of the line are monitored
      by photomultiplier tubes.   The photocurrents  from the photomultiplier
      tubes are processed and controlled by a computer system.  A background
      correction technique is required to compensate for variable background
      contribution to the determination of trace elements.  Background must be
      measured adjacent to analyte lines on samples during analysis.  The
      position selected for the  background intensity measurement, on either or
+A bibliography citing method references appears in paragraph 11 of the
 method.

~&
 CLP-M modified for the Contract Laboratory Program.

                                   D-18                                ILM01.0

-------
      both sides of the analytical line, will be determined by the complexity
      of the spectrum adjacent to the analyte line.  The position used must be
      free of spectral interference and reflect the same change in background
      intensity as occurs at the analyte wavelength measured.  Background
      correction is not required in cases of line broadening where a
      background correction measurement would actually degrade the analytical
      result.  The possibility of additional interferences named in 5.1 (and
      tests for their presence as described in 5.2) should also be recognized
      and appropriate corrections made.
                                                             «
3.     Definitions

3.1   Dissolved -- Those elements which will pass through a 0.45 urn membrane
      filter.

3.2   Suspended -- Those elements which are retained by a 0.45 urn membrane
      filter.

3.3   Total -- The concentration determined on an unfiltered sample following
      vigorous digestion.

3.4   Instrumental detection limits -- See Exhibit E.

3.5   Sensitivity -- The slope of the analytical curve, i.e.  functional
      relationship between emission intensity and concentration.

3.6   Instrument check standard -- A multi-element standard of known
      concentrations prepared by the analyst to monitor and verify instrument
      performance on a daily basis.  (See 7.6.1.)

3.7   Interference check sample - - A solution containing both interfering and
      analyte elements of known concentration that can be used to verify
      background and interelement correction factors.  (See 7.6.2.)

3.8   Quality control sample -- A solution obtained from an outside source
      having known concentration values to be used to verify the calibration
      standards.  (See 7.6.3.)

3.9   Calibration standards -- A series of known standard solutions used by
      the analyst for calibration of the instrument (i.e., preparation of the
      analytical curve).   (See 7.4.)

3.10  Linear dynamic range - - The concentration range over which the
      analytical curve remains linear as determined in Exhibit E.

3.11  Reagent blank -- A volume of deionized,  distilled water containing the
      same acid matrix as the calibration standards carried through the entire
      analytical scheme.   (See 7.5.2.)

3.12  Calibration blank -- A volume of deionized, distilled water acidified
      with HN03 and HC1.   (See 7.5.1.)
                                   D-19                               ILM01.0

-------
3.13  Method of standard addition -- The standard addition technique involves
      the use of the unknown and the unknown-plus-a-known amount of standard
      by adding known amounts of standard to one or more aliquots of the
      processed sample solution.

4.    Safety

4.1   The toxicity or carcinogenicity of each reagent used in this method has
      not been precisely defined; however, each chemical compound should be
      treated as a potential health hazard.   The laboratory is responsible for
      maintaining a current awareness file of OSHA regulations regarding the
      safe handling of the chemicals specified in this method.  A reference
      file of material handling data sheets should be made available to all
      personnel involved in the chemical analysis.

5.    Interferences

5.1   Several types of interference effects may contribute to inaccuracies in
      the determination of trace elements.  They can be summarized as follows:

      5.1.1   Spectral interferences can be categorized as 1) overl'ap of a
              spectral line from another element; 2) unresolved overlap of
              molecular band spectra; 3) background contribution from
              continuous or recombination phenomena; and 4) background
              contribution from stray light from the line emission of high
              concentration elements.  The first of these effects can be
              compensated by utilizing a computer correction of the raw data,
              requiring the monitoring and measurement of the interfering
              element.  The second effect may require selection of an
              alternate wavelength.  The third and fourth effects can usually
              be compensated by a background correction adjacent to the
              analyte line.  In addition, users of simultaneous multi-element
              instrumentation must assume the responsibility of verifying the
              absence of spectral interference from an element that could
              occur in a sample but for which there is no channel in the
              instrument array.

              Listed in Table 2 are some interference effects for the
              recommended wavelengths given in Table 1.  The data in Table 2
              are intended for use only as a rudimentary guide for the
              indication of potential spectral interferences.  For this
              purpose, linear relations between concentration and intensity
              for the analytes and the interferents can be assumed.  The
              interference information,  which was collected at the Ames
                        •Jf-jf
              Laboratory  , is expressed as analyte concentration equivalents
              (i.e., false analyte concentrations) arising from 100 mg/L of
              the interferent element.
  Ames Laboratory,  USDOE,  Iowa State University,  Ames,  Iowa 50011.

                                   D-20                                IIM01.0

-------
        The suggested use of this information is as follows:  Assume
        that arsenic (at 193.696 nm) is to be determined in a sample
        containing approximately 10 mg/L of of aluminum.  According to
        Table 2, 100 mg/L of aluminum would yield a false signal for
        arsenic equivalent to approximately 1.3 mg/L.  Therefore, 10
        mg/L of aluminum would result in a false signal for arsenic
        equivalent to approximately 0.13 mg/L.  The reader is cautioned
        that other analytical systems may exhibit somewhat different
        levels of interference than those shown in Table 2, and that
        the interference effects must be evaluated for, each individual
        system.  Only those interferents listed were investigated and
        the blank spaces in Table 2 indicate that measurable
        interferences were not observed from the interferent
        concentrations listed in Table 3.  Generally, interferences
        were discernible if they produced peaks or background shifts
        corresponding to 2-5% of the peaks generated by the analyte
        concentrations also listed in Table 3.

        At present, information on the listed silver and potassium
        wavelengths are not available but it has been reported that
        second order energy from the magnesium 383.231 nm wavelength
        interferes with the listed potassium line at 766.491 nm.

5.1.2   Physical interferences are generally considered to be effects
        associated with the sample nebulization and transport
        processes.  Such properties as change in viscosity and surface
        tension can cause significant inaccuracies especially in
        samples which may contain high dissolved solids and/or acid
        concentrations.  The use of a. peristaltic pump may lessen these
        interferences.  If these types of interferences are operative,
        they must be reduced by dilution of the sample and/ or
        utilization of standard addition techniques.  Another problem
        which can occur from high dissolved solids is salt buildup at
        the tip of the nebulizer.  This affects aerosol flow rate
        causing instrumental drift.

        Wetting the argon prior to nebulization, the use of a. tip
        washer, or sample dilution have been used to control this
        problem.  Also, it has been reported that better control of the
        argon flow rate improves instrument performance.  This is
        accomplished with the use of mass flow controllers.

5.1.3   Chemical interferences are characterized by molecular compound
        formation, ionization effects and solute vaporization effects.
        Normally these effects are not pronounced with the ICP
        technique, however, if observed they can be minimized by
        careful selection of operating conditions (that is, incident
        power, observation position, and so forth),  by buffering of the
        sample, by matrix matching, and by standard addition
        procedures.  These types of interferences can be highly
        dependent on matrix type and the specific analyte element.
                             D-21                               ILM01.0

-------
5.2  Prior to reporting concentration data for the  analyte elements,  the
     Contractor must analyze and report the results of the ICP Serial Dilution
     Analysis.   The ICP Serial Dilution Analysis  must be performed on a sample
     from each group of samples of a similar matrix type (i.e.,  water,  soil)
     and concentration (i.e.,  low,  medium) or for each Sample Delivery Group,
     whichever is more frequent.   Samples identified as field blanks  cannot be
     used for Serial Dilution Analysis.

      If the analyte concentration is sufficiently high (minimally a factor of
      50 above the instrumental detection limit in the original sample), the
      serial dilution (a five fold dilution) must then agree within 10% of the
      original determination after correction for dilution.  If the dilution
      analysis for one or more analytes is not within 10%, a chemical or
      physical interference effect must be suspected, and the data for all
      affected analytes in the samples received associated with that serial
      dilution must be flagged with an "E" on FORM IX-IN and FORM I-IN.

6.     Apparatus

6.1   Inductively Coupled Plasma-Atomic Emission Spectrometer.

      6.1.1   Computer controlled atomic  emission  spectrometer with
              background correction.

      6.1.2   Radio  frequency  generator.

      6.1.3   Argon  gas supply, welding  grade or better.

6.2   Operating conditions -- Because of the differences between various makes
      and models of satisfactory instruments, no detailed operating
      instructions can be provided.  Instead, the analyst should follow the
      instructions provided by the manufacturer of the particular instrument.
      Sensitivity, instrumental detection limit,  precision, linear dynamic
      range, and interference effects must be investigated and established for
      each individual analyte line on that particular instrument.  All
      measurements must be within the instrument linear ranee where correction
      factors are valid.  It is the responsibility of the analyst to verify
      that the instrument configuration and operating conditions used satisfy
      the analytical requirements and to maintain quality control data
      confirming instrument performance and analytical results.

7.     Reagents and Standards

7.1   Acids used in  the preparation of standards and for sample processing
      must be ultra-high purity grade or  equivalent.  Redistilled acids are
      acceptable.

      7.1.1    Acetic acid,  cone.   (sp gr 1.06).

      7.1.2    Hydrochloric  acid,  cone.   (sp gr  1.19).

      7.1.3    Hydrochloric  acid,  (1+1):   Add 500 mL cone.   HC1  (sp gr 1.19)
               to 400 mL deionized,  distilled water and dilute to 1 liter.
                                   D-22                                ILM01.0

-------
      7.1.4    Nitric  acid,  cone.   (sp  gr  1.41).

      7.1.5    Nitric  acid,  (1+1):  Add 500 mL  cone.   HNO3  (sp  gr  1.41)  to 400
               mL deionized,  distilled  water and dilute  to  1  liter.

7.2   Deionized, distilled water:  Prepare by passing distilled water through
      a mixed bed of cation and anion exchange resins.  Use deionized,
      distilled water for the preparation of all reagents,  calibration
      standards and as dilution water.   The purity of this water must be
      equivalent to ASTM Type II reagent water of Specification D 1193.

7.3   Standard stock solutions may be purchased or prepared from ultra high
      purity grade chemicals or metals.  All salts must be dried for 1 hour at
      105° unless otherwise specified.

      (CAUTION:  Many metal salts are extremely toxic and may be fatal if
      swallowed.  Wash hands thoroughly after handling.) Typical stock
      solution preparation procedures follow:

      7.3.1   Aluminum solution, stock, 1 mL - 100 ug Al:  Dissolved 0.100 g
              of aluminum metal in an acid mixture of 4 mL of  (1+1)' HC1 and 1
              mL of cone.  HNO-j in a beaker.  Warm gently to effect solution.
              When solution is complete, transfer quantitatively to a liter
              flask, add an additional 10 mL of (1+1) HC1 and  dilute to 1000
              mL with deionized, distilled water.

      7.3.2   Antimony solution stock,  1 mL - 100 ug Sb:  Dissolve 0.2669 g
              K(SbO)C4H^Og in deionized distilled water, add 10 mL  (1+1) HC1
              and dilute to 1000 mL with deionized, distilled water.

      7.3.3   Arsenic solution, stock,  1 mL - 100 ug As:  Dissolve 0.1320 g
              of AS20-J in 100 mL of deionized, distilled water containing 0.4
              g NaOH.  Acidify the solution with 2 mL cone.  HNO^ and dilute
              to 1,000 mL with deionized, distilled water.

      7.3.4   Barium solution, stock, 1 mL •• 100 ug Ba:  Dissolve 0.1516 g
              BaCl2 (dried at 250°C for 2 hrt.) in 10 mL deionized, distilled
              water with 1 mL (1+1) HC1.  Add 10.0 mL (1+1) HC1 and dilute to•
              1,000 mL with deionized,  distilled water.

      7.3.5   Beryllium solution, stock, 1 mL - 100 ug Be:  Do not dry.
              Dissolve 1.966 g BeSO^'4H20, in deionized, distilled water, add
              10.0 mL cone. HNOo and dilute to 1,000 mL with deionized,
              distilled water.

      7.3.6   Boron solution, stock, 1 mL - 100 ug B:  Do not  dry.  Dissolve
              0.5716 g anhydrous H-jBO-j in deionized, distilled water and
              dilute to 1,000 mL.  Use a reagent meeting ACS specifications,
              keep the bottle tightly stoppered and store in a desiccator to
              prevent the entrance of atmospheric moisture.
                                   D-23                               ILM01.0

-------
7.3.7    Cadmium solution,  stock, 1 mL - 100 ug Cd:  Dissolve  0.1142 g
         CdO in a minimum amount of (1+1) HNO-j.  Heat  to  increase rate
         of dissolution.  Add  10.0 mL cone. HNOo and dilute  to 1,000 mL
         with deionized,  distilled water.

7.3.8    Calcium solution,  stock, 1 mL - 100 ug Ca:  Suspend 0.2498  g
         CaC03 dried  at 180°C  for 1 h before weighing  in  deionized,
         distilled water  and dissolve cautiously with  a minimum amount
         of (1+1)  HN03.   Add 10.0 mL cone.  HN03 and dilute  to 1,000 mL
         with deionized,  distilled water.

7.3.9    Chromium solution, stock, 1 mL - 100 ug Cr:   Dissolve 0.1923 g
         of Cr02  in deionized, distilled water.  When  solution is
         complete acidify with 10 mL cone.  HN03 and dilute  to 1,000 mL
         with deionized,  distilled water.

7.3.10   Cobalt solution  stock, 1 mL - 100 ug Co:  Dissolve  0.1000 g of
         cobalt metal in  a minimum amount of (1+1) HN03.  Add  10.0 mL
         (1+1)  HC1 and dilute to 1,000 mL with deionized, distilled
         water.

7.3.11   Copper solution, stock, 1 mL - 100 ug Cu:  Dissolve 0.1252  g
         CuO in a minimum amount of (1+1) HNO^.  Add 10.0 mL cone.   HNO-j
         and dilute to 1,000 mL with deionized, distilled water.

7.3.12   Iron solution, stock, 1 mL - 100 ug Fe:  Dissolve 0.1430 g
         Fe203  in a warm  mixture of 20 mL (1+1) HC1 and 2 mL of cone.
         HNO3.   Cool, add an additional 5 mL of cone.  HNO-j  and dilute
         to 1,000 mL with deionized, distilled water.

7.3.13   Lead solution, stock, 1 mL - 100 ug Pb:  Dissolve 0.1599 g
         Pb(N03)2  in a minimum amount of (1+1) HN03.   Add 10.0 mL of
         cone.  HN03 and  dilute to 1,000 mL with deionized,  distilled
         water.

7.3.14   Magnesium solution, stock, 1 mL - 100 ug Mg:  Dissolve 0.1658 g
         MgO in a minimum amount of (1+1) HNO-,.  Add 10.0 mL cone.   HNO^
         and dilute to 1,000 mL with deionized, distilled water.

7.3.15   Manganese solution, stock, 1 mL - 100 ug Mn:  Dissolve 0.1000 g
         of manganese metal in the acid mixture, 10 mL cone.   HC1 and 1
         mL cone.  HN03,  and dilute to 1,000 mL with deionized,
         distilled water.

7.3.16   Molybdenum solution, stock, 1 mL - 100 ug Mo:  Dissolve  0.2043
         g  (NH4)2Mo04 in  deionized, distilled water and dilute to 1,000
         mL.

7.3.17   Nickel solution, stock, 1 mL -= 100 ug Ni:  Dissolve 0.1000  g of
         nickel metal in  10 mL hot cone.  HN03, cool and  dilute to 1,000
         mL with  deionized, distilled water.
                             D-24                                ILM01.0

-------
      7.3.18  Potassium solution, stock, 1 mL - 100 ug K:  Dissolve 0.1907 g
              KC1, dried at 110°C, in deionized, distilled water.  Dilute to
              1,000 mL.

      7.3.19  Selenium solution, stock, 1 mL - 100 ug Se:  Do not dry.
              Dissolve 0.1727 g l^SeO-j  (actual assay 94.6%) in deionized,
              distilled water and dilute to 1,000 mL.

      7.3.20  Silica solution, stock, 1 mL - 100 ug Si02:  Do not dry.
              Dissolve 0.4730 g Na2SiO-j'9H20 in deionized, distilled water.
              Add 10.0 mL cone.  HNO-j and dilute to 1,000 mL with deionized,
              distilled water.

      7.3.21  Silver solution, stock, 1 mL - 100 ug Ag:  Dissolve 0.1575 g
              AgN03 in 100 mL of deionized, distilled water and 10 mL cone.
              HN03.  Dilute to 1,000 mL with deionized, distilled water.

      7.3.22  Sodium solution, stock, 1 mL - 100 ug Na:  Dissolve 0.2542 g
              NaCl in  ieionized, distilled water.  Add 10.0 mL cone.  HNO-j
              and dilute to 1,000 mL with deionized, distilled water.

      7.3.23  Thallium solution, stock, 1 mL - 100 ug Tl:  Dissolve 0.1303 g
              TlNO-j in deionized, distilled water.  Add
              10.0 mL cone.  HNO^ and dilute to 1,000 mL with deionized,
              distilled water.

      7.3.24  Vanadium solution, stock, 1 mL - 100 ug V:  Dissolve 0.2297
              NH,VOo in a minimum amount of cone.  HNO^.  Heat to increase
              rate of dissolution.  Add 10.0 mL cone.  HNOo and dilute  to
              1,000 mL with deionized,  distilled water.

      7.3.25  Zinc solution, stock, 1 mL - 100 ug Zn:  Dissolve 0.1245  g ZnO
              in a minimum amount of dilute HNOo.  Add 10.0 mL cone. HNO-j and
              dilute to 1,000 mL with deionized, distilled water.

7.4   Mixed calibration standard solutions -- Prepare mixed calibration
      standard solutions by combining appropriate volumes of the stock
      solutions in volumetric flasks.   (See 7.4.1 thru 7.4.5.)  Add 2 mL of
      (1+1) HNO3 and 10 mL of (1+1)  HC1 and dilute to 100 mL with deionized,
      distilled water.   (See NOTE in 7.4.5) Prior to preparing the mixed
      standards, each stock solution should be analyzed separately to
      determine possible spectral interference or the presence  of impurities.
      Care should be taken when preparing the mixed standards that the
      elements are compatible and stable.   Transfer the mixed standard
      solutions to a FEP fluorocarbon or unused polyethylene bottle for
      storage.  Fresh mixed standards should be prepared as needed with the
      realization that concentration can change on aging.  Calibration
      standards must be initially verified using a quality control sample and
      monitored weekly for stability (see 7.6.3).  Although not specifically
      required,  some typical calibration standard combinations follow when
      using those specific wavelengths listed in Table 1.

      7.4.1   Mixed standard solution  I -- Manganese, beryllium,  cadmium,
              lead, and zinc.

                                   D-25                                ILMOL.O

-------
      7.4.2   Mixed standard solution II -- Barium, copper, iron, vanadium,
              and cobalt.

      7.4.3   Mixed standard solution III -- Molybdenum, silica, arsenic, and
              selenium.

      7.4.4   Mixed standard solution IV -- Calcium, sodium, potassium,
              aluminum, chromium and nickel.

      7.4.5   Mixed standard solution V -- Antimony, boron, aagnesium,
              silver, and thallium.

              NOTE:  If the addition of silver to the recommended acid
              combination results in an initial precipitation add 15 mL of
              deionized distilled water and warm the flask until the solution
              clears.  Cool and dilute to 100 mL with deionized, distilled
              water.  For this acid combination the silver concentration
              should be limited to 2 mg/L.  Silver under these conditions is
              stable in a tap water matrix for 30 days.  Higher
              concentrations of silver require additional HC1.

7.5   Two types  of blanks are  required for the analysis.   The  calibration
      blank (3.13)  is used in  establishing the analytical curve  while the
      reagent blank (preparation blank,  3.12)  is  used to correct for possible
      contamination resulting  from varying amounts of the acids  used in the
      sample processing.

      7.5.1   The calibration blank is prepared by diluting 2 mL of  (1+1)
              HN03 and 10 mL of (1+1) HC1 to 100 mL with deionized,  distilled
              water.  Prepare a sufficient quantity to be used to flush the
              system between standards and samples.

      7.5.2   The reagent blank (or preparation blank - See Exhibit  E) must
              contain all the reagents and in the same volumes as used in the
              processing of the samples.  The reagent blank must be  carried
              through the complete procedure and contain the same acid
              concentration in the final solution as the sample solution used
              for analysis.

7.6   In addition the calibration standards,  an instrument check standard
      (3.6), an  interference check sample (3.7)  and a quality  control sample
      (3.8) are  also required  for the analyses.

      7.6.1   The instrument check standard for continuing calibration
              verification is prepared by the analyst by combining compatible
              elements at a concentration equivalent to the mid-points of
              their respective calibration curves.  (See 10.1.3.)

      7.6.2   The interference check sample is prepared by the analyst, or
              obtained from EPA if available (Exhibit E).
                                   D-26                                 ILM01.0

-------
       7.6.3    The quality control sample for the initial calibration
               verification should be prepared in the same acid matrix as the
               calibration standards and in accordance with the instructions
               provided by the supplier.  EPA will either supply a quality
               control sample or information where one of equal quality can be
               procured.  (See 10.1.1.)

8.     Procedure

8.1    Set up instrument with proper  operating parameters established in
       Section 6.2.  The instrument must be allowed to become  thermally stable
       before beginning.  This usually requires at least SO min.  of operation
       prior to calibration.

8.2    Initiate appropriate operating configuration of computer.

8.3    Profile and calibrate instrument  according to instrument manufacturer's
       recommended procedures, using mixed calibration standard solutions such
       as those described in Section 7.4.  Flush the system with the
       calibration blank (7.5.1) between each standard.  (NOTE:  For boron
       concentrations greater than 500 ug/L extended flush times of 1 to 2
      minutes may be required.)

8.4    Begin the sample  run flushing the system with the calibration blank
       solution (7.5.1) between each sample.  (See NOTE in 8.3.)  Analyze the
       instrument check standard (7.6.1) and the calibration blank (7.5.1) each
       10 analytical samples.

8.5   A minimum of two replicate exposures are required for standardization
      and all QC and sample analyses.  The average result of  the multiple
      exposures for the standardization and all QC and sample analyses shall
      be used.

9.    Calculation

9.1   Reagent blanks (preparation blanks) should be treated as specified in
      Exhibit E.

9.2   If dilutions were performed, the appropriate factor must be applied to
      sample values.

9.3   Units must be clearly specified.

10.   Quality Control (Instrumental)

10.1  Quality control must be performed as specified in Exhibit E.

11.  Bibliography

     1.  Winge, R.K., V.J. Peterson, and  V.A. Fassel,  "Inductively Coupled
         Plasma-Atomic  Emission Spectroscopy Prominent Lines," EPA-600/4-79-
         017.
                                   D-27                                ILM01.0

-------
2.   Winefordner,  J.D.,  "Trace Analysis:  Spectroscopic Methods for
     Elements,"  Chemical Analysis, Vol. 46, pp. 41-42.

3.   Handbook  for  Analytical Quality Control  in Water and Wastewater
     Laboratories,  EPA-600/4-79-019.

4.   Garbarino,  J.R.  and Taylor,  K.E.,  "An Inductively-Coupled Plasma
     Atomic  Emission  Spectrometric Method for Routine Water Quality
     Testing," Applied Spectroscopy 33, No. 3(1979).

5.   "Methods  for  Chemical Analysis of  Water  and Wastes," EPA-600/4-79-
     020.

6.   Annual  Book of ASTM Standards, Part 31.

7.   "Carcinogens  - Working With  Carcinogens," Department of Health,
     Education,  and Welfare, Public Health Service, Center for Disease
     Control,  National Institute  for Occupational Safety and Health,
     Publication No.  77-206, Aug. 1977.

8.   "OSHA Safety  and Health Standards, General Industry," (29 CFR 1910),
     Occupational  Safety and Health Administration, OSHA 2206, (Revised,
     January 1976).

9.   "Safety in  Academic Chemistry Laboratories, American Chemical
     Society Publications, Committee on Chemical Safety, 3rd Edition,
     1979.

10.  "Inductively  Coupled Plasma-Atomic Emission Spectrometric Method of
     Trace Elements Analysis of Water and Waste", Method 200.7 modified
     by CLP  Inorganic Data/Protocol Review Committee; original method by
     Theodore  D. Martin, EMSL/Cincinnati.
                              D-28                                ILM01.0

-------
               TABLE 1  -  RECOMMENDED WAVELENGTHS(2)  AND ESTIMATED
                         INSTRUMENTAL DETECTION LIMITS
Element
Aluminum
Antimony
Arsenic
Barium
Beryllium
Boron
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Potassium
Selenium
Silica (Si02)
Silver
Sodium
Thallium
Vanadium
Zinc
Wavelength, nm(l)
308.215
206.833
193.696
455.403
313.042
249.773
226.502
317.933
267.716
228.616
324.754
259.940
220.353
279.079
257.610
202.030
231.604
766.491
196.026
288.158
328.068
588.995
190.864
292.402
213.856
Estimated Detection
Limit, ug/L(2)
45
32
53
2
0.3
5
4
10
7
7
6
7
42
30
2
8
15
See (3)
75
58
7
29
40
8
2
(1)   The  wavelengths  listed are  recommended because of their  sensitivity and
     overall  acceptance.   Other  wavelengths may be  substituted  if they can
     provide  the needed  sensitivity and are treated with  the  same corrective
     techniques  for spectral interference.   (See 5.1.1).  The  use  of  alternate
     wavelengths must be reported (in nm) with  the  sample data.

(2)   The  estimated instrumental  detection limits as shown are taken  from
     "Inductively Coupled Plasma-Atomic Emission Spectroscopy-Prominent
     Lines,"  EPA-600/4-79-017.   They are given  as a. guide for an  instrumental
     limit.   The actual  method detection limits are sample dependent and may
     vary as  the sample  matrix varies.

(3)   Highly dependent on operating conditions and plasma  position.
                                    D-29                                ILMOL.O

-------







s
0
K
rf
f-<
O
2
H
CQ
t-t
(Z
<

^^
>1
\
o
^_^

cn
EH
2
Cd
iJ
<

H
D
O
U ,4
U
2 >
o ca
M ^
E-i
gt?
OS \
E-i CJ)
2 S
u
O 0
2 0
O rH
o
H 93
EH EH
>i
J EH
<~f*
<,
2
< cn
£H
Cn 2
O W
K
W W
-1 r
\ — 1 PH
a, t£
2 U
< EH
X 2
U M


•
CN
U
ij
CQ
EH



























-P
C
0)
r-l
0)





•H
EH


•H
2



C
s



en
s





0)
f
M-l




3
cj




J-l
CJ




«J
CJ




r-l
<









C







Q)
-P
>i
r-l
(0
c
<
in
•** <• rH
. . .
<-H O t-H

in
1 CM 1
1 • 1


1 1 1
1 1 1

t-H
CM 1 1
• 1 1
0



1 1 1
1 1 1




CO
1 0 1
II
• 1
o




1 1 1
1 1 1


If
I en T
i • •
CN O




1 1 1
1 1 1



r-
1 •* co
1 • •
O rH





in co vo
t-H co cn
CN CO VO
co vo n
o o en
to CN <-H




g >>
3 C O
C 0 -H
-H g C
g -H Q)
3 -P w
rH C M
< < <
in
1 0 1
1 • 1
0

Tf
1 0 1
1 • 1
o

1 1 1
1 1 1


1 1 1
1 1 1




1 1 1
1 1 1




CM
1 1 CO
Ii —
1 •
0




1 1 1
1 1 1



1 1 1
1 1 1





1 1 1
1 1 1



«*
1 1 0
1 1 •
o





CO CN CO
o •* r-
•cr o t-*
tn co en
in t-H ^r
TJ- ro CN



g
3
-rH
g rH
3 rH C
-H >i O
S-i S-i SH
tO  C
,.* 0)
i— 1 O rH
O-H 0)
S 2 cn
rH
O 1 1
• 1 1
O

CO
1 0 1
1 • 1
o

1 t t
1 1 1


1 1 1
1 1 i




1 1 1
1 1 1





1 1 1
il 1
1 1




1 1 1
1 1 1


r-
0 1 1
• i i
o




1 1 1
1 1 1



o
1 1 CO
1 1 •
o





co in •<*
in cn vo
rH en co
co co o
co co en
CN in  N




o
.
—->
O
y
,J
r— 1



























O
co
1
Q






































-------
TABLE 3.  INTERFERENT AND ANALYTE ELEMENTAL CONCENTRATIONS USED
           FOR INTERFERENCE MEASUREMENTS IN TABLE 2
Analytes
Al
As
B
Ba
Be
Ca
Cd
Co
Cr
Cu
Fe
Mg
Mn
Mo
Na
Ni
Pb
Sb
Se
Si
Tl
V
Zn
(mg/L)
10
10
10
1
1
1
10
1
1
1
1
1
1
10
10
10
10
10
10
1
10
1
10
Interferents
Al
Ca
Cr
Cu
Fe
Mg
Mn
Ni
Ti
V












(mg/L)
1000
1000
200
200
1000
1000
200
200
200
200












                              D-31                                ILM01.0

-------
           PART B -  ATOMIC ABSORPTION METHODS.  FURNACE TECHNIQUE'1'
      Ana1vte/Method                                          Page  No.

      Antimony -  Method 204.2 CLP-M*                           D-33
      Arsenic - Method 206.2 CLP-M                             D-34
      Beryllium - Method 210.2 CLP-M                           D-36
      Cadmium - Method 213.2 CLP-M                             D-37
      Chromium -  Method 218.2 CLP-M                            D-38
      Lead - Method 239.2 CLP-M                                D-39
      Selenium -  Method 270.2 CLP-M                            D-41
      Silver - Method 2/2.2 CLP-M                              D-43
      Thallium -  Method 279.2 CLP-M                            D-44
+From "Methods for Chemical Analysis of Water and Wastes" (EPA-600/4-79-
 020), Metals-4, as modified for use in the Contract Laboratory Program).
"CLP-M modified for the Contract Laboratory Program.
                                   D-32                                ILM01.0

-------
                                  ANTIMONY

         Method 204.2 CLP-M  (Atomic Absorption, Furnace Technique)
Optimum Concentration Range:  20-300 ug/L
Approximate Detection Limit;:  3 ug/L
Preparation of Standard Solution
1.   Stock solution:  Carefully weigh 2.7426 g of antimony potassium
     tartrate (analytical reagent grade)  and dissolve in deionized distilled
     water.   Dilute to 1 Liter with deionized water.  1 mL - 1 mg Sb (1000
     ag/L).
2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.   These solutions are also to be used
     for "standard additions".
3.   The calibration standards  must be prepared using the same type of acid
     and at  the same concentration as will result in the sample to be
     analyzed after sample preparation.
Instrument Parameters (General)
1.   Drying  Time and Temp:  30  sec @ 125°C.
2.   Ashing  Time and Temp:  30  sec @ 800°C.
3.   Atomizing Time and Temp:   10 sec <§ 2700°C.
4.   Purge Gas Atmosphere:  Argon
5.   Wavelength:  217.6 run
6.   Other operating parameters should be set as specified by the particular
     ins trument manufacture r.
1.   The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100,  based on the use of a. 20 uL injection,  contin-
     uous flow purge gas and non-pyrolytic graphite and are to be used as
     guidelines only.   Smaller size furnace devices or those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter time periods than the above recommended
     sett-ings.

2.   The use of background correction is required.

3.   Nitrogen may also be used as the purge gas.

4.   If chloride concentration presents a matrix problem or causes a loss
     previous to atomization, add an excess 5 mg of ammonium nitrate to the
     furnace and ash using a ramp accessory or with incremental steps until
     the recommended ashing temperature is reached.

5.   For every sample analyzed, verification is necessary to determine that
     method of standard addition is not required (see Exhibit E).

6.   If method of standard addition is required, follow the procedure given
     in Exhibit E.
 CLP-M modified for the Contract Laboratory Program.
                                   D-33                                ILM01.0

-------
                                   ARSENIC

          Method 206.2 CLP-M  (Atomic Absorption, Furnace Technique)
 Optimum Concentration Range:   5-100 ug/L
 Approximate Detection Limit:   1 ug/L

 Preparation of Standard Solution

 1.    Stock solution:   Dissolve  1.320 g of arsenic  trioxide, AS202
      (analytical  reagent  grade)  in 100 mL of  deionized  distilled water
      containing 4 g NaOH.   Acidify the solution  with  20 mL cone. HNO-, and
      dilute to 1  Liter.   1  mL - 1  mg As (1000 mg/1).

 2.    Nickel Nitrate Solution, 5%:   Dissolve 24.780 g  of ACS reagent  grade
      Ni(NO-j)2 '6^0 in deionized distilled water  and make up to  100 mL.

 3.    Nickel Nitrate Solution, 1%:   Dilute 20  mL  of the  5% nickel nitrate to
      100 mL with  deionized  distilled water.

 4.    Working Arsenic  Solution:   Prepare dilutions of  the stock  solution to
      be used as calibration standards  at the  time of  aaal^jis.    Withdraw
      appropriate  aliquots of  the stock solution, add  1 mL of cone.  HNCU,  2
      mL of  30% H202 and 2 mL  of  the  5%  nickel nitrate solution.   Dilute to
      100 mL with  deionized  distilled water.

 Sample Preparation

 1.   Add 100 uL of the 5% nickel nitrate  solution to 5 mL of the digested
      sample.  The sample is now ready  for injection into the furnace.
 Instrument  Parameters  (General)

 1.   Drying Time and Temp:  30 sec @ 125°C.
 2.   Ashing Time and Temp:  30 sec @ 1100°C.
 3.   Atomizing Time and Temp:   10 sec @ 2700°C.
 4.   Purge Gas Atmosphere:  Argon
 5.   Wavelength:  193.7 nm
 6.   Other operating parameters  should be set as specified by  the particular
     instrument manufacturer.
Notes
     The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100, based on the use of a 20 uL injection,  purge gas
     interrupt and non-pyrolytic graphite.   Smaller size furnace devices or
     those employing faster rates of atomization can be operated using lower
     atomization temperatures for shorter time periods  than the above
     recommended settings.

     The use of background  correction is required.  Background correction
     made by the deuterium  arc method does  not adequately compensate for
     high levels of certain interferents (ie., Al,  Fe).   If conditions occur
     where significant interference is suspected,  the lab must switch to an
     alternate wavelength or take other appropriate actions to compensate
     for the interference effects.
 CLP-M modified for the Contract Laboratory Program.

                                   D-34                               ILM01.0

-------
3.   For every sample analyzed,  verification is necessary to determine that
     method of standard addition is not required (see Exhibit E).
4.   If method of standard addition is required,  follow the  procedure given
     in Exhibit E).

5.   The use of the Electrodeless Discharge  Lamps (EDL) for  the light source
     is recommended.
                                  D-35                                ILM01.0

-------
                                  BERYLLIUM

         Method 210.2 CLP-M   (Atomic Absorption, Furnace Technique)
Optimum Concentration Range:  1-30 ug/L
Approximate Detection Limit:  0.2 ug/L
Preparation of Standard Solution

1.   Stock solution:  Dissolve 11.6586g of beryllium sulfate,  BeSO^,  in
     deionized distilled water containing 2 mL concentrated nitric acid and
     dilute to 1 Liter.  1 mL - 1 mg Be (1000 mg/L).

2.   Prepare dilutions of che stock solution to be used as calibration
     standards at the time of analysis.   These solutions are also to be
     used for "standard additions".

3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

1.   Drying Time and Temp:  30 sec @ 125°C.
2.   Ashing Time and Temp:  30 sec @ 1000°C.
3.   Atomizing Time and Temp:  10 sec @ 2800°C.
4.   Purge Gas Atmosphere:  Argon
5.   Wavelength:  234.9 nin
6.   The operating parameters should be set as specified by the particular
     instrument manufacturer.
1.    The above concentration values and instrument conditions are for a
      Perkin-Elmer HGA-2100, based on the use of a 20 uL injection, con-
      tinuous flow purge gas and non-pyrolytic graphite and are to be used
      as guidelines only.  Smaller size furnace devices or those employing
      faster rates of atomization can be operated using lower atomization
      temperatures for shorter time periods than the above recommended
      settings.
2.    The use of background correction is required.

3.    Because of possible chemical interaction, nitrogen  should  not  be used
      as a purge gas.

4.    For  every sample  analyzed,  verification  is  necessary to  determine
      that method of standard addition is not required (see Exhibit E)
5.   If method of standard addition is required,  follow the procedure given
     in Exhibit E.
 CLP-M  modified  for  the Contract Laboratory Program.
                                   D-36                                ILM01.0

-------
                                   CADMIUM

         Method 213.2 CLP-M   (Atomic Absorption,  Furnace Technique)
Optimum Concentration Range:  0.5-10 ug/L
Approximate Detection Limit:  0.1 ug/L

Preparation of Standard Solution

1.   Stock solution:  Carefully weigh 2.282g of cadmium sulfate, 3 CdSO^'  8
     HoO (analytical reagent grade) and dissolve in deionized distilled
     water.  Make up to 1 Liter with deionized distilled water.  1 mL — 1 mg
     Cd (1000 mg/L).

2.   Ammonium Phosphate solution (40%):  Dissolve 40 grams of ammonium
     phosphate, (NH^)2HPO^ (analytical reagent grade) in deionized distilled
     water and dilute to 100 mL.

3.   Prepare dilutions of stock cadmium solution to be used as calibration
     standards at the time of analysis.  To each 100 mL of standard and
     sample alike add 2.0 mL of the ammonium phosphate solution.  The
     calibration standards must be prepared using the same type of acid and
     at the same concentration as will result in the sample to be analyzed
     after sample preparation.

Instrument Parameters (General)

1.   Drying Time and Temp:  30 sec @ 125°C.
2.   Ashing Time and Temp:  30 sec @ 500°C.
3.   Atomizing Time and Temp:  10 sec @ 1900°C.
4.   Purge Gas Atmosphere:  Argon
5.   Wavelength:  228.8 nm
6.   The operating parameters should be set as specified by the particular
     instrument manufacturer.

Notes

1.   The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100, based on the use of a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite and are to be used
     as guidelines only.   Smaller size furnace devices or those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter time periods than the above recommended
     settings.

2.   The use of background correction is required.

3.   Contamination from  the work area is critical in cadmium analysis.   Use
     pipette tips which  are free of cadmium.

4.   For every sample analyzed,  verification is necessary to determine  that
     method of standard  addition is not required (see Exhibit E) .

5.   If method of standard addition is  required,  follow the  procedure given
     in Exhibit E.
 CLP-M modified for the Contract Laboratory Program.
                                   D-37                               ILM01.0

-------
                                   CHROMIUM

          Method 218.2  CLP-M   (Atomic Absorption,  Furnace Technique)
Optimum  Concentration  Range:   5-100  ug/L
Approximate  Detection  Limit:   1  ug/L
Preparation  of  Standard  Solution
1.   Stock solution:   Prepare as described under Part C methods, AA Flame
     Technique.
2.   Calcium Nitrate solution:  Dissolve 11.8 grams of calcium nitrate,
     CaCNO^^'^HoO  (analytical reagent grade) in deionized distilled water
     and dilute to  100 mL.    1 mL - 20 mg Ca.
3.   Prepare dilutions of the stock  chromium solution to be used as
     calibration standards at the  time of analysis.  The calibration
     standards must be prepared using the same type of acid and at the same
     concentration  as will result  in the sample to be analyzed after sample
     preparation.  To  each 100 mL  of standard and sample alike, add 1 mL of
     30% ^02 and 1 mL of che calcium nitrate solution.
Instrument Parameters  (General)
1.   Drying Time and Temp:  30 sec @ 125°C.
2.   Ashing Time and Temp:  30 sec @ 1000°C.
3.   Atomizing Time and Temp:  10  sec @ 2700°C.
4.   Purge Gas Atmosphere:  Argon
5.   Wavelength:  357.9 nm
6.   Other operating parameters should be set as specified by the particular
     instrument manufacturer.
1.   The above concentration values and instrument conditions are for a
     Perkin Elmer HGA-2100, based on the use of a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphits and are to be used
     as guidelines only.

2.   Hydrogen peroxide is added to the acidified solution to convert all
     chromium to the trivalent state.     Calcium is added to a level' above
     200 mg/L where its suppressive effect becomes constant up to 1000 mg/L.

3.   Background correction is required.

4.   Nitrogen should not be used as a purge gas because of possible CN band
     interference.

5.   Pipette tips have been reported to be a possible source  of
     contamination.

6.   For every sample analyzed,  verification is necessary to  determine that
     method of standard addition is not required (see Exhibit E).

7.   If method of standard addition is required,  follow the procedure  given
     in Exhibit E.
 CLP-M modified for the Contract Laboratory Program.
                                   D-38                               ILM01.0

-------
                                    LEAD

         Method 239.2 CLP-M*  (Atomic Absorption, Furnace Technique)

Optimum Concentration Range:  5-100 ug/L
Approximate Detection Limic:  1 ug/L
Preparation of Standard Solution
1.   Stock solution:  Carefully weigh 1.599 g of lead nitrate,  Pb(N03)2
     (analytical reagent grade), and dissolve in deionized distilled water.
     When solution is complete,  acidify with 10 mL redistilled HNO^ and
     dilute to 1 Liter with deionized distilled water.  1 mL - 1 rag Pb
     (lOOOmg/L).
2.   Lanthanum Nitrate' solution:  Dissolve 58.64 g of ACS reagent grade
     La203 in 100 mL cone.  HN03  and dilute to 1000 mL with deionized
     distilled water.  1 mL - 50 mg La.
3.   Working Lead solution:  Prepare dilutions of stock lead solution to  be
     used as calibration standards at the time of analysis.   The calibration
     standards must be prepared  using the same type of acid and at the same
     concentration as will  result in the sample to be analyzed after sample
     preparation.   To each  100 mL of diluted standard add 10 mL of the
     lanthanum nitrate solution.
Sample Preparation
1.   To each 100 mL of prepared  sample solution add 10 mL of the lanthanum
     nitrate solution.
Instrument Parameters (General)

1.   Drying Time and Temp:   30 sec @ 125°C.
2.   Ashing Time and Temp:   30 sec @ 500°C.
3.   Atomizing Time and Temp:   10 sec @ 2700°C.
4.   Purge Gas Atmosphere:   Argon
5.   Wavelength:   283.3 nm
6.   Other operating parameters  should be set as specified by the particular
     ins trument manufacture r.
     The above concentration values and instrument conditions  are  for  a
     Perkin-Elmer HGA-2100,  based on the use  of a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite  and  are  to be  used
     as guidelines only.   Smaller size furnace  devices  or  those  employing
     faster rates of atomization can be operated using  lower atomization
     temperatures for shorter time periods  than the above  recommended
     settings.

     The use  of background correction is required.
 CLP-M modified for the Contract Laboratory Program.
                                   D-39                               ILM01.0

-------
3.   Greater sensitivity can be acheived using the  217.0 nm line,  but the
     optimum concentration range is  reduced.   The use  of a lead
     electrodeless discharge lamp at this lower wavelength has  been found to
     be advantageous.   Also a lower  atomization temperature (2400°C)  may be
     preferred.

4.   To suppress sulfate interference (up to  1500 ppm)  lanthanum is added as
     the nitrate to both samples and calibration standards.   (Atomic
     Absorption Newsletter Vol. 15,  No.  3,  p.  71, May-June 1976).

5.   Since glassware contamination is a  severe problem in lead  analysis,  all
     glassware should be cleaned immediately  prior  to  use,  and  once cleaned,
     should not be open to the atmosphere except when  necessary.

6.   For every sample analyzed, verification  is necessary to  determine  that
     method of standard addition is  not  required (see  Exhibit E).

7.   If method of standard addition  is required, follow the procedure given
     in Exhibit E.
                                   D-40                               ILM01.0

-------
                                  SELENIUM

         Method  270.2  CLP-M   (Atomic  Absorption,  Furnace Technique)

Optimum  Concentration  Range:   5-100 ug/L
Approximate Detection  Limit:   2 ug/L

Preparation of Standard  Solution

1.   Stock Selenium solution:  Dissolve 0.3453 g of selenous acid  (actual
     assay 94.6% J^SeO-j) in deionized distilled water and make up  to 200 mL.
     1 mL - 1 mg Se (1000 nig/L) .

2.   Nickel Nitrate solution,  5%:  Dissolve 24.780 g of ACS reagent grade
     NiCNO-j^. 6H20 in  deionized distilled water and make up to 100 mL.

3.   Nickel Nitrate solution,  1%:  Dilute 20 mL of the 5% nickel nitrate to
     100 mL with deionized distilled water.

4.   Working Selenium  solution:  Prepare dilutions of the stock solution to
     be used as calibration standards at the time of analysis.  The
     calibration standard" must be prepared using the samp, type of acid and
     at the same concentration as will result in the sample to be analyzed
     after sample preparation.  Withdraw appropriate aliquots of the stock
     solution, add 1 mL of cone.  HNO^, 2 mL of 30% 1^02 and 2 mL of the 5%
     nickel nitrate solution.  Dilute to 100 mL with deionized distilled
     water.

Sample Preparation

1.   Add 100 uL of the 5% nickel nitrate solution to 5 mL of the digested
     sample.   The sample is now ready for injection into the furnace.
Instrument Parameters

1.   Drying Time and Temp:   30 sec @ 125°C.
2.   Charring Time and Temp:   30 sec @ 1200°C.
3.   Atomizing Time and Temp:  10 sec @ 2700°C.
4.   Purge Gas Atmosphere:   Argon
5.   Wavelength:   196.0 nm
6.   Other operating parameters should be set as  specified by the particular
     ins trument manufacture r.
Notes

1.   The  above concentration values  and instrument conditions are for  a
     Perkin-Elmer HGA-2100,  based on the use of a 20 uL injection,  purge  gas
     interrupt and non-pyrolytic  graphite and are to be used as  guidelines
     only.   Smaller size  furnace  devices or those employing faster  rates  of
     atomization can be operated using lower atomization tempera-  tures for
     shorter  time periods than the above recommended settings.

2.   The  use  of background  correction is required.     Background correction
     made by  the  deuterium  arc method does not adequately compensate for
     high levels  of certain interferents (i.e., Al,  Fe).   If conditions
 CLP-M modified for the Contract Laboratory Program.
                                                                      ILM01.0

-------
     occur where significant interference is suspected,  the lab must switch
     to an alternate wavelength or take other appropriate actions to
     compensate for the interference effects.

3.   Selenium analysis suffers interference from chlorides (>800 mg/L)  and
     sulfate (>200 mg/L).   For the analysis of industrial effluents  and
     samples with concentrations of sulfate from 200 to  2000 mg/L, both
     samples and standards should be prepared to contain 1% nickel.
4.   For every sample analyzed, verification is necessary to determine  that
     method of standard addition is not required (see Exhibit E).

5.   If method of standard addition is required, follow  the procedure given
     in Exhibit E.

6.   The use of the Electrodeless Discharge Lamp (EDL) for the light source
     is recommended.
                                  D-42                                ILM01.0

-------
                                   SILVER

         Method 272.2 CLP-M*  (Atomic Absorption, Furnace Technique)
Optimum Concentration Range:  1-25 ug/L
Approximate Detection Limit.:  0.2 ug/L
Preparation of Standard Solution
1.   Stock solution:  Dissolve 1.575 g of AgN03 (analytical reagent grade)
     in deionized distilled water.  Add 10 mL of concentrated HNO^ and make
     up to 1 Liter.  1 mL - 1 mg Ag (1000 mg/L).
2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  These solutions are also to be used
     for "standard additions".
3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be
     analyzed after sample preparation.
Instrument Parameters (General)
1.   Drying Time and Temp:  30 sec @ 125°C.
2.   Astr'ng Time and Temp:  30 sec @ 400°C.
3.   Atomizing Time and Temp:  10 sec @ 2700°C.
4.   Purge Gas Atmosphere:  Argon
5.   Wavelength:  328,1 nni
6.   Other operating parameters should be set as specified by the particular
     instrument manufacturer.
1.   The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100, based on the use of a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite and are to be used
     as guidelines only.  Smaller size furnace devices or those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter time periods than the above recommended
     settings.

2.   The use of background correction is required.

3.   The use of halide acids should be avoided.
4.   If absorption to container walls or formation of AgCl is suspected,  see
     Part G, AA methods Flame Technique.
5.   For every sample analyzed, verification is necessary to determine that
     method of standard addition is not required (see Exhibit E) .
6.   If method of standard addition is required, follow the procedure given
     in Exhibit E.
 CLP-M modified  for the Contract Laboratory  Program.
                                   D-43                                ILM01.0

-------
                                  THALLIUM

         Method 279.2 CLP-M   (Atomic Absorption, Furnace Technique)
Optimum Concentration Range:  5-100 ug/L
Approximate Detection Limit:  1 ug/L

Preparation of Standard Solution

1.   Stock solution:  Dissolve 1.303g of thallium nitrate,  TINO-j  (analytical
     reagent grade) in deionized distilled water.  Add 10 mL of concentrated
     nitric acid and dilute to 1 Liter with deionized distilled water.   1 mL
     - 1 mg Tl (1000 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  These solutions are also to be used
     for "standard additions".

3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be.
     analyzed after sample preparation.

Instrument Parameters (General)

1.   Drying Time and Temp:  30 sec @ 125°C.
2.   Ashing Time and Temp:  30 sec @ 400°C.
3.   Atomizing Time and Temp:  10 sec @ 2400°C.
4.   Purge Gas Atmosphere:  Argon
5.   Wavelength:  276.8 nm
6.   Other operating'parameters should be set as specified by the particular
     instrument manufacturer.

Notes

1.    The  above  concentration values  and instrument  conditions are  for a
      Perkin-Elmer  HGA-2100,  based  on  the  use   of  a  20  uL  injection,
      continuous  flow  purge  gas  and non-pyrolytic graphite  and are  to  be
      used  as  guidelines only.    Smaller   size  furnace  devices  or  those
      employing  faster  rates  of atontization  can be  operated  using  lower
      atomization  tempera Cures  for shorter  time  periods  than the  above
      recommended settings.

2.    The use of background correction is required.
3.    Nitrogen may also be used as the purge gas.

4.    For  every  sample  analyzed,  verification  is necessary to  determine
      that method of standard addition is not required (see Exhibit E).

5.    If  method  of  standard  addition is  required,   follow  the  procedure
      given in Exhibit E.
 CLP-M modified for the Contract Laboratory Program.

                                   D-44                               ILM01.0

-------
             PART  C  - ATOMIC ABSORPTION METHODS...  FLAME TECHNIQUE'1'
      Analyte/Method                                          Page No,

      Calcium  - Method  215.1 CLP-M*                             D-46
      Magnesium - Method  242.1 CLP-M                            D-47
      Potassium - Method  258.1 CLP-M                            D-48
      Sodium - Method 273.1 CLP-M                               D-49
 From "Interim Methods for the Sampling and Analysis of Priority Pollutants
 in Sediments and Fish Tissue", USEPA EMSL, Cincinnati, Ohio, August 1977,
 Revised October 1980, as modified for use in the Contract Laboratory
 Program.
~CLP-M modified for the Contract Laboratory Program.
                                   °-45                               ILM01.0

-------
                                  CALCIUM

          Method 215.1 CLP-M."  (Atomic Absorption,  Flame Technique)

Optimum Concentration Range:  0.2-7 mg/L using a wavelength of 422.7 nm
Sensitivity:  0.08 mg/L
Detection Limit:  0.01 mg/L
Preparation of Standard Solution
1.   Stock Solution:   Suspend 1.250 g of CaCO-j (analytical reagent grade),
     dried at 180°C for 1 hour before weighing,  in deionized distilled water
     and dissolve cautiously with a mimimum of dilute HC1.  Dilute to 1000
     mL with deionized distilled water.   1  mL - 0.5 mg Ca (500 mg/L).
2.   Lanthanum chloride solution:  Dissolve 29 g of La2°3. slowly and in
     small portions,  in 250 mL cone.   HC1 (Caution:  Reaction is violent)
     and dilute to 500 mL with deionized distilled water.
3.   Prepare dilutions of the stock calcium solutions to be used as
     calibration standards at the time of analysis.  To each 10 mL of
     calibration standard -ad sample alike  add 1.0 mL of the lanthanum
     chloride solution,  i.e.,  20 mL of standard or sample + 2 mL LaCl-^ - 22
     mL.
Instrumental Parameters (General)
1.   Calcium hollow cathode lamp
2.   Wavelength:  422.7 nm
3.   Fuel:  Acetylene
4.   Oxidant:  Air
5.   Type of flame:   Reducing
1.   Phosphate,  sulfate and aluminum interfere  but are  masked by the
     addition of lanthanum.  Because low calcium values result if the pH of
     the sample  is above 7, both standards  and  samples  are prepared in
     dilute hydrochloric acid solution.   Concentrations of magnesium greater
     than. 1000 mg/L also cause low calcium  values.   Concentrations of up to
     500 mg/L each of sodium, potassium and nitrate cause no interference.

2.   Anionic chemical interferences can be  expected if  lanthanum is not used
     in samples  and standards.
3.   The nitrous oxide-acetylene flame will provide two to five times
     greater sensitivity and freedom from chemical inteferences.   lonization
     interferences should be controlled by  adding a large amount of alkali
     to the sample and standards.   The analysis appears to be free from
     chemical suppressions in the nitrous oxide-acetylene flame.   (Atomic
     Absorption  Newsletter 14, 29 [1975]).
                j
4.   The 239.9 nm line may also be used. This  line has a relative
     sensitivity of 120.
^CLP-M modified for the Contract Laboratory Program.

                                   D-46                                ILM01.0

-------
                                 MAGNESIUM

          Method 242.1 CLP-M  (Atomic Absorption,  Flame Technique)

Optimum Concentration Range:  0.02-0.5 mg/L using a wavelength of 285.2 nm
Sensitivity:  0.007 mg/L
Detection Limit:  0.001 mg/L
Preparation of Standard Solution
1.   Stock Solution:   Dissolve 0.829  g of magnesium oxide,  MgO (analytical
     reagent grade),  in 10 mL of redistilled HNO-j  and dilute to 1 liter with
     deionized distilled water.   1  mL - 0.50 mg Mg (500 mg/L).
2.   Lanthanum chloride solution:   Dissolve  29  g of La2°3»  slowly and in
     small portions in 250 mL concentrated HC1  (Caution:  Reaction  is
     violent), and dilute to 500 mL with deionized distilled water.
3.   Prepare dilutions of the stock magnesium solution to be used as
     calibration standards at the time of analysis.   To each 10 mL  volume of
     calibration standard and sample  alike add  1.0 mL of the lanthanum
     chloride solution,  i.e., 20 mL of standard or sample + 2 mL LaCl-j - 2±
     mL.
Instrumental Parameters (General)
1.   Magnesium hollow cathode lamp
2.   Wavelength:   285.2 nm
3.   Fuel:  Acetylene
4.   Oxidant:  Air
5.   Type of flame:  Oxidizing
     The interference caused by aluminum at  concentrations  greater  than 2
     mg/L is masked by addition of lanthanum.   Sodium,  potassium and calcium
     cause no interference at concentrations less  than  400  mg/L.

2.   The following line may also be used:  202.5 nm Relative  Sensitivity 25.

3.   To cover the range of magnesium values  normally observed in surface
     waters (0.1-20 mg/L),  it is suggested that either  the  202.5 nm line be
     used or the  burner head be rotated.   A  90° rotation  of the  burner  head
     will produce approximately one-eighth the  normal sensitivity.
"CLP-K modified for the Contract Laboratory Program.

                                   D-47                               ILM01.0

-------
                                  POTASSIUM

          Method 258.1 CLP-M* (Atomic Absorption, Flame Technique)

Optimum Concentration Range:  0.1-2 mg/L using a wavelength of 766.5 nm
Sensitivity:  0.04 mg/L
Detection Limit:  0.01 mg/L

Preparation of Standard Solution

1.   Stock Solution:   Dissolve 0.1907 g of KC1 (analytical reagent grade),
     dried at 110°C,  in deionized distilled water and make up  to 1 liter.  1
     mL - 0.10 mg K (100 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  The calibration standards should be
     prepared using the same type of acid and at the same concentration as
     will result in the sample to be analyzed either directly  or after
     processing.

Instrumental Parameters {General)

1.   Potassium hollow cathode lamp
2.   Wavelength:   766.5 nin
3.   Fuel:   Acetylene
4.   Oxidant:   Air
5.   Type of flame:   Slightly oxidizing

Notes

1.   In air-acetylene or other high temperature flames (>2800°C),  potassium
     can experience partial ionization which indirectly affects  absorption
     sensitivity.  The presence of other alkali salts in the sample can
     reduce this ionization and thereby enhance analytical results.   The
     ionization suppressive effect of sodium is small if the ratio of Na to
     K is under 10.   Any enhancement due to sodium can be stabilized by
     adding excess sodium (1000 ug/mL) to both sample and standard
     solutions.  If more stringent control of ionization is required,  the
     addition of cesium should be considered.   Reagent blanks  must be
     analyzed to correct for potassium impurities in the buffer  zone.
2.   The 404.4 nm line may also  be used.   This line  has a relative
     sensitivity of 500.

3.   To cover the range of potassium values normally observed  in surface
     waters (0.1-20 mg/L), it is suggested that the  burner head  be rotated.
     A 90  rotation of the burner head provides approximately  one-eighth the
     normal sensitivity.
"CLP-M modified for the Contract Laboratory Program.

                                   D-48                                ILM01.0

-------
                                    SODIUM

          Method  273.1 CLP-M   (Atomic Absorption,  Flame Technique)

Optimum Concentration Range:   0.03-1 mg/L using  a  wavelength of 589.6 nm
Sensitivity:   0.015 mg/L
Detection Limit:   0.002 mg/L

Preparation  of Standard Solutions

1.   Stock Solution:  Dissolve 2.542 g of NaCl (analytical reagent grade),
     dried at  140°C, in deionized distilled water and make up to 1 liter. 1
     mL - 1 mg Na  (1000 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at  the time of analysis.   The calibration standards should be
     prepared using the same type of acid and at the same concentration as  •
     will result in the sample to be analyzed either directly or after
     processing.

Instrumental Parameters (General)
1.   Sodium hollow cathode lamp
2.   Wavelength:  589.6 nra
3.   Fuel:   Acetylene
4.   Oxidant:  Air
5.   Type of flame:  Oxidizing
1.   The 330.2 nm resonance line of sodium,  which has a relative sensitivity
     of 185, provides a convenient way to avoid the need to dilute more
     concentrated solutions of sodium.

2.   Low-temperature flames increase sensitivity by reducing the extent of
     ionization of this easily ionized metal.   lonization may also be
     controlled by adding potassium (1000 mg/L)  to both standards  and
     samples.
CCLP-M  modified for the Contract Laboratory Program.

                                   D'A9                                1LM01.0

-------
           PART D -  COLD VAPOR METHODS  FOR MERCURY ANALYSIS4
Method                                                             Page No.

Mercury Analysis in Water by Manual Cold Vapor Technique            D-50
Method 245.1 CLP-M*

Mercury Analysis in Water by Automated Cold Vapor Technique         D-58
Method 245.2 CLP-M

Mercury Analysis in Soil/Sediment by Manual Cold Vapor Technique    D-64
Method 245.5 CLP-M
 A bibliography citing method references follows  each method.
^
 CLP-M modified for the Contract Laboratory Program.

                                   D-50                               ILM01.0

-------
          MERCURY ANALYSIS IN WATER BY MANUAL COLD VAPOR TECHNIQUE

                                  MERCURY
             Method  245.1  CLP-M*  (Manual Cold Vapor  Technique)
1.     Scope and Application

1.1   In addition to inorganic forms of mercury, organic mercurials may
      also be present.  These organo-mercury compounds will not respond to
      the cold vapor atomic absorption technique unless they are first
      broken down and converted to mercuric ions.   Potassium permanganate
      oxidizes many of these compounds, but recent studies have shown that
      a number of organic mercurials,  including phenyl mercuric acetate and
      methyl mercuric chloride, are only partially oxidized by this
      reagent.  Potassium persulfate has been found to give approximately
      100% recovery when used as the oxidant with these compounds.
      Therefore, a persulfate oxidation step following the addition of the
      permanganate has been included to insure that organo-mercury
      compounds, if present, will be oxidized to the mercuric ion before
      measurement.  A heat step is required for methyl mercuric chloride
     . when present in or spiked to a natural system.

1.2   The range of the method may be varied through instrument and/or
      recorder expansion.  Using a 100 mL sample,  a detection limit of 0.2
      ug Hg/L can be achieved (See 10.2).

2.     Summary of Method

2.1   The flameless AA procedure is a physical method based on the
      absorption of radiation at 253.7 rim by mercury vapor.  Organic
      mercury compounds are oxidized and the mercury is reduced to the
      elemental state and aerated from solution in a closed system.  The
      mercury vapor passes through a cell positioned in the light path of
      an atomic absorption spectrophotometer.   Absorbance (peak height) is
      measured as a function of mercury concentration and recorded in the
      usual manner.

3.     Sample Handling and Preservation

3.1   Until more conclusive data are obtained, samples are preserved by
      acidification with nitric acid to a pH of 2 or lower immediately at
      the time of coll.iction (Exhibit D, Section II).

4.     Interference

4.1   Possible interference from sulfide is eliminated by the addition of
      ^otassium'permanganate.  Concentrations as high as 20 mg/1 of sulfide
      as sodium sulfide do not interfere with the recovery of added
      inorganic mercury from distilled water (Exhibit D, Section II).
NCLP-M modified for the Contract Laboratory Program.

                                   D-51                                ILM01.0

-------
4.2   Copper has also been reported Co interfere; however, copper
      concentrations as high as 10 mg/L had no effect on recovery of
      mercury from spiked samples.

4.3   Sea waters, brines and industrial effluents high in chlorides require
      additional permanganate (as much as 25 mL).  During the oxidation
      step, chlorides are converted to free chlorine which will also absorb
      radiation of 253 nm.  Care must be taken to assure that free chlorine
      is absent before the mercury is reduced and swept into the cell.
      This may be accomplished by using an excess of hydroxylamine sulfate
      reagent (25 mL) .  Both inorganic and organic mercury spikes have been
      quantitatively recovered from the sea water using this technique.

5.     Apparatus

5.1   Atomic Absorption Spectrophotometer:  (See Note 1) Any atomic
      absorption unit having an open sample presentation area in which to
      mount the absorption cell is suitable.  Instrument settings
      recommended by the particular manufacturer should be followed.

      NOTE 1:  Instruments designed specifically for the measurement of
      mercury using the cold vapor technique are commercially available and
      may be substituted for the atomic absorption spectrophotometer.

5.2   Mercury Hollow Cathode Lamp:   Westinghouse WL-22847, argon filled, or
      equivalent.

5.3   Recorder:  Any multi-range variable speed recorder that is compatible
      with the UV detection system is suitable.

5.4   Absorption Cell:   Standard spectrophotometer cells 10 cm long, having
      quartz end windows may be used.  Suitable cells may be constructed
      from plexiglass tubing,  1" O.D. X 4-1/2".   The ends are ground
      perpendicular to the longitudinal axis and quartz windows (1"
      diameter X 1/16"  thickness) are cemented in place.

      The" cell is strapped to a burner for support and aligned in the  light
      beam by use of two 2" by 2" cards.   One inch diameter holes are  cut
      in the middle of each card; the cards are  then placed over each  end
      of the cell.   The cell is then positioned and adjusted vertically and
      horizontally to find the maximum transmittance.

5.5   Air Pump:   Any peristaltic pump capable of delivering 1 liter of air
      per minute may be used.   A Masterflex pump with electronic speed
      control has been found to be  satisfactory.

5.6   Flowmeter:  Capable of measuring an air flow of 1 liter per minute.

5.7   Aeration Tubing:   A straight glass  frit having a coarse porosity.
      Tygon tubing is used for passage of the mercury vapor from the sample
      bottle to the absorption cell and return.
                                   D-52                                ILM01.0

-------
5.8   Drying Tube:  6" X 3/4" diameter  tube containing  20 g of magnesium
      perchlorate  (see Note  2).  The  apparatus  is  assembled as shown in
      Figure 1.

      NOTE 2:  In  place of the magnesium perchlorate drying tube, a small
      reading lamp with 60U'  bulb ma'"  be used  to prevent condensation of
      moisture inside the cell.  The  lamp is  positioned to shine on the
      absorption cell maintaining the air temperature in the cell about
      10°C above ambient.

6.    Reagents

6.1   Sulfuric Acid, Cone:   Reagent grade.

      6.1.1    Sulfuric acid, 0.5 N:   Dilute 14.0 mL of  cone.  sulfuric acid
               to 1.0  liter.

6.2   Nitric Acid,  Cone:   Reagent grade of low mercury content (see Note
      3).

      NOTE 3:  If a high reagent blank is obtained, it may be necessary to
      distill the nitric acid.

6.3   Stannous Sulfate:   Add 25 g stannous sulfate to 250 mL of 0.5 N
      sulfuric acid.  This mixture is a suspension and should be stirred
      continuously during use.  (Stannous chloride may be used in place of
      stannous sulfate.)

6.4   Sodium Chloride-Hyroxylamine Sulfate Solution:  Dissolve 12 g of
      sodium chloride and 12 g of hydroxylamine sulfate in distilled water
      and dilute to 100 mL.   (Hydroxylamine hydrochloride may be used in
      place of hydroxylamine sulfate.)

6.5   Potassium Permanganate:  5% solution,  w/v.  Dissolve 5 g of potassium
      permanganate in 100 mL of distilled water.

6.6   Potassium Persulfate:   5% solution,  w/v.  Dissolve 5 g of potassium
      persulfate in 100 mL of distilled water.

    6.7   Stock Mercury Solution:   Dissolve 0.1354 g of mercuric chloride
          in 75 mL  of distilled water.  Add 10 mL of cone,  nitric  acid  and
          adjust the volume  to 100.0 mL.  1  mL — 1  mg Hg.
                                   D-53                                ILM01.0

-------
         Figure 1.  Apparatus for Flameless Mercury Determination
f
u
AIR PUMP
» f 1
r 1 I
" 7 DESICCANT .
r
ARcnp
\ U* BUBBLER CE
k 	
^
V
M.
1
PTION
LL
•)
)
1

Iv
D-*
SAMPLE  SOLUTION
IN BOD  BOTTLE
SCRUBBER
CONTAINING
A MERCURY
ABSORBING
MEDIA
                             D-54
                                                            ILM01.0

-------
6.8   Working Mercury Solution:  Make successive dilutions of the stock
      mercury solution to obtain a working standard containing 0.1 ug per
      mL.  This working standard and the dilutions of the stock mercury
      solution should be prepared fresh daily.  Acidity of the working
      standard should be maintained at 0.15% nitric acid.  This acid should
      be added to the flask as needed before the addition of the aliquot.

7.     Calibration

7.1   Transfer 0, 0.5, 1.0, 5.0 and 10.0 mL aliquots of the working mercury
      solution containing 0 to 1.0 ug of mercury to a series of 300 mL BOD
      bottles.  Add enough distilled water to each bottle to make a total
      volume of 100 mL.  Mix thoroughly and add 5 mL of cone,  sulfuric
      acid (6.1) and 2.5 mL of cone,  nitric acid (6.2) to each bottle.
      Add 15 mL of KMnO^ (6.5) solution to each bottle and allow to stand
      at least 15 minutes.  Add 8 mL of potassium persulfate (6.6) to each
      bottle and heat for 2 hours in a water bath maintained at 95°C.
      Alternatively, cover the BOD bottles with foil and heat in an
      autoclave for 15 minutes at 120°C and 15 Ibs.  Cool and add 6 mL of
      sodium chloride-hydroxylamine sulfate solution (6.4) to reduce the
      excess permanganate.  When the solution has been decolorized wait 30
      seconds, add 5 mL of the stannous sulfate solution (6.3) and
      immediately attach the bottle to the aeration apparatus forming a
      closed system.  At this point the sample is allowed to stand quietly
      without manual agitation.

      The circulating pump, which has previously been adjusted to a rate of
      1 liter per minute, is allowed to run continuously (see Note 4).  The
      absorbance will increase and reach maximum within 30 seconds.  As
      soon as the recorder pen levels off, approximately 1 minute, open the
      bypass valve and continue the aeration until the absorbance returns
      to its minimum value (see Note 5).  Close the bypass valve, remove
      the stopper and frit from the BOD bottle and continue the aeration.
      Proceed with the standards and construct a standard curve by plotting
      peak height versus micrograms of mercury.

      NOTE 4:  An open system where the mercury vapor is passed through the
      absorption cell only once may be used instead of the closed system.

      NOTE 5:  Because of the toxic nature of mercury vapor precaution must
      be taken to avoid its inhalation.  Therefore, a bypass has been
      included in the system to either vent the mercury vapor into an
      exhaust hood or pass th<_ vapor through some absorbing media,  such as:
      a)    equal volumes of 0.1 M KMnO*,  and 10% HoSO-   or
      b) 0.25% iodine in a 3% a KI solution.  A specially treated charcoal
      that will adsorb mercury vapor is available.
                                   D-55                                ILM01.0

-------
8.     Procedure

8.1   Transfer 100 mL, or an aliquot diluted to 100 mL, containing not more
      than 1.0 ug of mercury, to a 300 mL BOD bottle.  Add 5 mL of sulfuric
      acid (6.1) and 2.5 mL of cone.  nitric acid (6.2) mixing after each
      additon.  Add 15 mL of potassium permanganate solution (6.5) to each
      sample bottle (see Note 6).   For sewage samples additional
      permanganate may be required.   Shake and add additional portions of
      potassium permanganate solution, if necessary,  until the purple color
      persists for at least 15 minutes.  Add 8 mL of potassium persulfate
      (6.6) to each bottle and heat for 2 hours in a water bath at 95°C.

      NOTE 6:   The same amount of KMnO^ added to the samples should be
      present in standards and blanks.

      Cool and add 6 mL of sodium chloride-hydroxylamine sulfate (6.4) to
      reduce the excess permanganate (see Note 7).   Purge the head space  in
      the BOD bottle for at least 1 minute and add 5 mL of Stannous Sulfate
      (6.3) and immediately attach the bottle to the aeration apparatus.
      Continue as described under Calibration.

      NOTE 7:   Add reductant in 6  mL increments until KMnO^ is completely
      reduced.

9.     Calculation

9.1   Determine the-peak height of the unknown from the chart and read the
      mercury value from the standard curve.

9.2   Calculate the mercury concentration in the sample by the formula:

                       ug Hg in                 1,000
          ug Hg/L  -    aliquot    x
                                        volume of aliquot  in mL

10.    Appendix

10.1  If additional sensitivity is required,  a 200 mL sample with recorder
      expansion may be used provided the instrument does not produce undue
      noise.   Using a Coleman MAS-50 with a drying tube of magnesium
      perchlorate and a variable recorder,  2  mv was set to read full scale.
      With these conditions,  and distilled water solutions of mercuric
      chloride at concentrations of 0.15,  0.10,  0.05 and 0.025 ug/L the
      standard deviations were ±0.027,  +0.0006,  +0.01 and  +0.004.   Percent
      recoveries at these levels were 107,  83,  84 and 96%,  respectively.

10.2  Directions for the disposal  of mercury-containing wastes are given  in
      ASTM Standards, Part 31,  "Water",  p.  349,  Method D3223 (1976).
                                   D-56                                ILM01.0

-------
Bibliography

1.   Kopp,  J.F.,  Longbottom,  M.C.  and Lobring,  L.B.  "Cold Vapor Method  for
     Determining  Mercury",  AWWA,  vol.  64,  p.  20,  Jan.  1972.

2.   Annual Book  of ASTM Standards,  Part  31,  "Water",  Standard D3223-73, p.
     343 (1976).

3.   Standard Methods  for the  Examination  of  Water and Wastewater 14th
     Edition,  p.  156 (1975).
                                  D-57                                ILM01.0

-------
        MERCURY ANALYSIS  IN WATER  BY AUTOMATED COLD VAPOR TECHNIQUE

                                   MERCURY
            Method 245.2 CLP-M* (Automated Cold Vapor Technique)


1.     Scope and Application

1.1   The working range is 0.2 to 20.0 ug Hg/L.

2.     Summary of Method

2.1   The flameless AA procedure is a physical method based on the
      absorption of radiation at 253.7 nm by mercury vapor.   The  mercury is
      reduced to the elemental state and aerated from solution.  The
      mercury vapor passes through a cell positioned in the light path of
      an atomic absorption spectrophoCometer.  Absorbance (peak height) is
      measured as a function of mercury concentration and recorded in the
      usual manner.

2.2   In addition to inorganic forms of mercury,  organic mercurials may
      also be present.   These organo-mercury compounds will not respond to
      the flameless atomic absorption technique unless they are first
      broken down and converted to mercuric ions.  Potassium permanganate
      oxidizes many of these compounds, but recent studies have shown that
      a number of organic mercurials, including phenyl mercuric acetate and
      methyl mercuric chloride, are only partially oxidized by this
      reagent.  Potassium persulfate has been found to give approximately
      100% recovery when used as the oxidant with these compounds.
      Therefore,  an automated persulfate oxidation step following the
      automated addition of the permanganate has  been included to insure
      that organo-mercury compounds, if present,  will be oxidized to the
      mercuric ion before measurement.

3.     Sample Handling and Preservation

3.1   Until more conclusive data are obtained, samples are preserved by
      acidification with nitric acid to a pH of 2 or lower immediately at
      the time of collection (Exhibit D, Section II).

4.     Interferences (see NOTE 1>

4.1   Some sea waters and waste-waters  high in chlorides have shown a
      positive interference,  probably due to the  formation of free
      chlorine.

4.2   Formation of a heavy precipitate, in some wastewaters  and effluents,
      has been reported upon addition of concentrated sulfuric acid.   If
      this is encountered, the problem sample cannot be analyzed  by this
      method.
 'CLP-M modified for the Contract Laboratory Program.

                                   D-58                               ILM01.0

-------
4.3   Samples containing solids must be blended and then mixed while being
      sampled if total mercury values are to be reported.

      NOTE 1:  All o£ the ^bove interferences can be overcome by use of the
      Manual Mercury method.

5.     Apparatus

5.1   Technicon Auto Analyzer or equivalent instrumentation consisting of:

      5.1.1   Sampler  II with provision for  sample mixing.

      5.1.2   Manifold.

      5.1.3   Proportioning Pump  II  or  III.

      5.1.4   High  temperature heating  bath  with  two distillation coils
              (Technicon Part #116-0163)  in  series.

5.2   Vapor-liquid separator (Figure 1).

5.3   Absorption cell, 100 mm long, 10 mm diameter with quartz windows.

5.4   Atomic Absorption Spectrophotometer (see Note 2):  Any atomic
      absorption unit having an open sample presentation area in which to
      mount the absorption cell is suitable.   Instrument settings
      recommended by the particular manufacturer should be followed.

      NOTE 2:  Instruments designed specifically for the measurement of
      mercury using the cold vapor technique are commercially available and
      may be substituted for the atomic absorption spectrophotometer.

5.5   Mercury Hollow Cathode Lamp:  Westinghouse WL-22847, argon filled,  or
      equivalent.

5.6   Recorder:   Any multi-range variable speed recorder that is compatible
      with the UV detection system is suitable.

6.     Reagents

6.1   Sulfuric Acid, Cone:   Reagent grade

      6.1.1   Sulfuric acid, 2 N:  Dilute 56 raL of cone, sulfuric acid to 1
              liter with distilled water.

      6.1.2   Sulfuric acid, 10%:  Dilute 100 mL cone,  sulfuric acid to 1
              liter with distilled water.

6.2   Nitric acid,  Cone:   Reagent grade  of low mercury  content.

      6.2.1.  Nitric Acid,  0.5% Wash Solution:  Dilute 5 mL of concentrated
              nitric acid  to 1 liter with distilled water.
                                                                      ILM01.0

-------
 6.3   Stannous Sulfate  (See Note 3):  Add 50 g stannous sulfate to 500 mL
      of 2 N  sulfuric acid (6.1.1).  This mixture is a suspension and
      should  be stirred continuously during use.

      NOTE 3:  Stannous chloride may be used in place of stannous sulfate.

 6.4   Sodium  Chloride-Hydroxylamine Sulfate (See Note 4) Solution:
      Dissolve 30 g of sodium chloride and 30 g of hydroxylamine sulfate in
      distilled water to 1 liter.

      NOTE 4:  Hydroxylaraine hydrochloride may be used in place of
      hydroxylamine sulfate.

 6.5   Potassium Permanganate:  0.5% solution,  w/v.  Dissolve 5 g of
      potassium permanganate in 1 liter of distilled water.

 6.6   Potassium Permanganate, 0.1 N:  Dissolve 3.16 g of potassium
      permanganate in distilled water and dilute to 1 liter.

 6.7   Potassium Persulfate:   0.5% solution, w/v.  Dissolve 5 g potassium
      persulfate in I liter of distilled water.

 6.8   Stock Mercury Solution:  Dissolve 0.1354 g of mercuric chloride in 75
      mL of distilled water.   Add 10 mL of cone, nitric acid and adjust the
      volume  to 100.0 mL.  1.0 mL - 1.0 mg Hg.

 6.9   Working Mercury Solution:  Make successive dilutions of the stock
      mercury solution (6.8)  to obtain a working standard containing 0.1 ug
      per mL.  This working standard and the dilutions of the stock mercury
      solution should be prepared fresh daily.   Acidity of the working
      standard should be maintained at 0.15% nitric acid.   This acid should
      be added to the flask as needed before the addition of the aliquot.
      From this solution prepare standards containing 0.2,  0.5,  1.0,  2.0,
      5.0, 10.0, 15.0 and 20.0 ug Hg/L-

 6.10  Air Scrubber Solution:   Mix equal volumes of 0.1 N potassium
      permanganate (6.6) and 10% sulfuric acid (6.1.2).

 7.    Procedure (See Note 5)

 7.1   Set up manifold as shown in Figure 2.

 7.2   Feeding all the reagents through the system with acid wash solution
      (6.2.1)  through the sample line,  adjust heating bath  to 105°C.

7.3   Turn on atomic absorption spectrophotometer,  adjust  instrument
      settings as recommended by the manufacturer,  align absorption cell  in
      light path for maximum  transmittance and  place  heat  lamp  directly
      over absorption cell.

7.4   Arrange working mercury standards  from 0.2  to 20.0 ug Hg/L in sampler
      and start sampling.   Complete loading of  sample  tray  with  unknown
      samples.


                                   D-60                                ILM01.0

-------
 7.5   Prepare  standard curve by plotting peak height  of processed  standards
      against  concentration values.  Determine concentration of samples by
      comparing sample peak height with standard curve.

 7.6   After the analysis is complete put all lines  except the t^SO^  line in
      distilled water to wash out system.  After flushing, wash out  the
      I^SO^ line.  Also flush the coils in the high temperature heating
      bath by  pumping stannous sulfate (6.3) through  the sample lines
      followed by distilled vater.  This will prevent build-up of  oxides of
      manganese.

      NOTE 5:  Because of che toxic nature of mercury vapor, precaution
      must be  taken to avoid its inhalation.  Venting the mercury  vapor
      into an  exhaust hood or passing the vapor through some absorbing
      media such as:  a) equal volumes of 0.1 N KMn04(6.6) and 10% I^SO^
      (6.1.2), or b)    0.25% iodine in a 3% KI solution, is recommended.
      A specially treated charcoal that will absorb mercury vapor  is also
      available.

 Bibliography

 1.   Wallace R.A.,  Fulkerson,  W.,  Shults,  W.D.,  and Lyon,  U.S.,  "Mercury in
     the Environment-The Human Element",  Oak Ridge National Laboratory,
     ORNL/NSF-EP-1  p.  31,  (January,  1971).

 2.   Hatch,  W.R.  and Ott,  W.L.,  "Determination of Sub-Microgram Quantities
     of Mercury by  Atomic Absorption Specrophotometry".   Anal.  Chem.  40,
     2085 (1968).

 3.   Brandenberger,  H.  and Bader,  H. ,  "The Determination of Nanogram Levels
     of Mercury in  Solution by a  Flameless Atomic Absorption Technique",
     Atomic Absorption Newsletter 6,  101  (1967).

 4.   Brandenberger,  H.  and Bader,  H.,  "The Determination of Mercury by
     Flameless Atomic  Absorption  II,  A Static Vapor Method",  Atomic
    Absorption Newsletter 7,53 (1968).

 5.   Goulden,  P.O.  and Afghan,  B.K.  "An Automated Method for Determining
     Mercury in Water",  Technicon, Adv.  in Auto.  Analy.  2, p. 317  (1970).

 6.   "Interim Methods  for  the  Sampling and  Analysis  of Priority Pollutants
     in  Sediments and  Fish  Tissue," USEPA,  Environmental Monitoring and
     Support  Laboratory,  Cincinnati,  Ohio,  August  1977,  revised October
     1980.

7.   Op.  cit.  (#1), Methods  245.1  or  245.2.
                                   D-61                                ILM01.0

-------
AIR  AND
SOLUTION
ft*
                O.4 cm ID
                               7/25 T
O.7 cm 10
                                              AIR
                                              OUT
                            Hem
                                                 SOLUTION
                                                  OUT
             Figure 1.  Vapor liquid separator
                          D-62
                                                           ILM01.0

-------

C
0
«-
3:

L
«r
^
<-*
O
•^.
3:
8
o
c»
«-»
3
c- :

O
o

*
c o
c- C
*" «*«
2 O
_ o

3
J

I
0
C"
«->
31
3.

I

c
o
I-
f%
2

8
1

1
0
o«
«n
i
a.

J
u
<:
c
1
c
c*
y.
c.

g

o
O*
r^
ex.

1

f—
O
c^
f*»
3t
a.

C
1
«^»
1
s
O
c-
<->
3
a.

J
~j
o
«/t
c
«n
*<
O
«
o
r>#
>-


u:
«•
6
^^i
z ^-«
»-» c-»
O
«r\
«w
C.


O
c
•^
0 0
^^ c*
>- 3
«

1
-^
O
«/>
*x
a:
c
o
^
*r
r««
Cs»
Or
«
-_
•»i.i
1
f
c
«;

- — >
J
5"
£
kC
W\
D
Nrf
•4

•

                            w I
                            o-
  3
  O

< O
                                                                            oc
                                                                            O
                                                                           •-- c
                                                                             O
                                                                           TJ —•

                                                                        c
                                                                        w4
                                                                        C
                                                                        O
                                                                        O
                                                                        ex
                                                                        O
                   Figure 2.  Mercury Manifold AA-1
                                 D-63
                                                                       ILM01.0

-------
      MERCURY ANALYSIS  IN  SOIL/SEDIMENT BY MANUAL COLD VAPOR TECHNIQUE

                           MERCURY  (in Sediments)
              Method 245.5  CLP-M* (Manual  Cold Vapor Technique)

 1.    Scope and Application

 1.1   This procedure measures total mercury (organic and inorganic)  in
      soils, sediments, bottom deposits and sludge  type materials
                                                            I
 1.2   The range of  the  method is 0.2 to 5 ug/g.  The range may be extended
      above or below the normal range by increasing or decreasing sample
      size or through instrument and recorder control

 2.    Summary of Method

 2.1   A weighed portion of the sample is acid digested for 2 minutes at
      95°C, followed by oxidation with potassium permanganate and potassium
      persulfate.  Mercury in the digested sample is then measured by the
      conventional cold vapor technique

 2.2   An alternate digestion involving the use of an autoclave is described
      in (c.2)

 3.    Sample Handling and Preservation

 3.1   Because of the extreme sensitivity of the analytical procedure and
      the omnipresence  of mercury,  care must be taken to avoid extraneous
      contamination.  Sampling devices and sample containers should be
      ascertained to be free of mercury;  the sample should not be exposed
      to any condition  in the laboratory that may result in contact or air-
      borne mercury contamination

 3.2   Refrigerate solid samples at 4°C (±2°)  upon receipt until analysis
      (see Exhibit D,  Section II).

 3.3   The i- -mple should be  analyzed without drying.   A separate percent
      solics determination  is required,  (Part F).

4.     Interferences

4.1   The same types of interferences that may occur in water  samples are
      also possible with sediments,  i.e.,  sulfides,  high copper,  high
      chlorides,  etc.

4.2   Samples containing high concentrations  of oxidizable organic
      materials,  as evidenced by high chemical oxygen demand values,  may
      not be completely oxidized by this procedure.   When thisoccurs, the
      recovery of organic mercury will be  low.   The  problem can be
      eliminated by reducing the v.-eight of the original  sample  or by
 CLP-M modified for the Contract Laboratory Program.
                                   D-64                               ILM01.0

-------
       increasing the  amount  of potassium  persulfate  (and  consequently
       stannous  chloride)  used  in  the  digestion.

 5.     Apparatus

 5.1    Atomic Absorption Spectrophotometer (see Note  1):   Any atomic
       absorption unit having an open  sample presentation  area in which  to
       mount the absorption cell is suitable.  Instrument  settings
       recommended by the  particular manufacturer should be followed

       NOTE 1:   Instruments designed specifically for the  measurement  of
       mercury using the cold vapor technique are commercially available and
       may be substituted  for the atomic absorption Spectrophotometer

 5.2    Mercury Hollow Cathode Lamp:  Westinghouse WL-22847, argon filled, or
       equivalent

 5.3    Recorder:   Any multi-range variable speed recorder  that is compatible
       with the  UV detection  system is suitable.

 5.4    Absorption  Cell:  Standard Spectrophotometer cells  10 cm long", having
       quartz end  windows  may be used.  Suitable cells many be constructed
       from pexiglass tubing, 1" O.D.  X 4-1/2".  The ends are ground
       perpendicular to the longitudinal axis and quartz windows (1"
       diameter  X  1/16" thickness) are cemented in place.  Gas inlet and
       outlet ports (also  of  plexiglass but 1/4" O.D.) are attached
       approximately 1/2"  from each end.  The cell is strapped to a burner
       for support and aligned in the light beam to give the maximum
       transmittance.   Two 2" X  2" cards with one inch diameter holes may be
       placed over each end of the cell to assist in positioning the cell
       for maximum transmittance.

 5.5    Air Pump:  Any peristaltic pump capable of delivering 1 liter of  air
       per minute may be used.  A Masterflex pump with electronic speed
       control has been found to be satisfatory.  (Regulated compressed air
       can be used in an open one-pass system.)

 5.6    Flowmeter:  Capable of measuring an air flow of 1 liter per minute

 5.7   Aeration Tubing:   Tygon tubing is used for passage of the mercury
      vapor from the  sample bottle to the absorption cell and return.
       Straight glass  tubing terminating in a coarse porous frit is used for
       sparging air into the sample

5.8   Drying Tube:  6"  X 3/4" diameter tube containing 20 g of magnesium
      perchlorate (see Note 2).

      NOTE 2:      In  place of the  magnesium perchlorate drying tube,  a
      small reading lamp with 60W  bulb may be used to prevent condensation
      of moisture inside  the  cell.   The lamp is positioned to shine on the
      absorption cell  maintaining  the  air  temperature in the  cell about
      10°C above ambient.
                                   D-65                                ILM01.0

-------
6.     Reagents

6.1   Sulfuric acid,  cone.:   Reagent grade of low mercury content

6.2   Nitric acid,  cone.:   Reagent grade of low mercury content

6.3   Stannous Sulfate:   Add 25 g stannous sulfate to 250 mL of 0.5 N
      sulfuric acid (6.2).   This mixture is a suspension and should be
      stirred continuously during use
                                                            c
6.4   Sodium Chloride-Hydroxylamine Sulfate (See Note 3) Solution:
      Dissolve 12 g of sodium chloride and 12 g of hydroxylamine sulfate in
      distilled water and dilute to 100 mL

      NOTE 3:     A 10%  solution of stannous chloride may be substituted
      for (6.3) and hydroxylamine hydrochloride may be used in place of
      hydroxylamine sulfate  in (6.4)

6.5   Potassium Permanganate:   5% solution, w/v.   Dissolve 5 g of potassium
      permanganate in 100 mL of distilled water

6.6   Potassium Persulfate:   5% solution, w/v.   Dissolve 5 g of potassium
      persulfate in 100  mL of distilled water

6.7   Stock Mercury Solution:   Dissolve 0.1354 g of mercuric chloride in 75
      mL of distilled water.   Add mL of cone,  nitric acid and adjust the
      volume to 100.0 mL.   1.0 - 1.0 mg Hg

6.8   Working Mercury Solution:  Make successive dilutions of the stock
      mercury solution (6.7) to obtain a working standard containing 0.1
      ug/mL.  This working standard and the dilution of the stock mercury
      solutions should be prepared fresh daily.  Acidity of the working
      standard should be maintained at 0.15% nitric acid.  This acid should
      be added to the flask as needed before the addition of the aliquot

7.     Calibration

7.1   Transfer 0, 0.5, 1.0,  5.0 and 10 mL aliquots of the working mercury
      solutions (6.8) containing 0 to 1.0 ug of mercury to a series of 300
      mL BOD bottles. Add enough distilled water to each bottle to make a
      total volume of 10 mL.   Add 5 mL of cone.  H2SO^ (6.1) and 2.5 mL of
      cone.  HNO3 (6.-2)  and heat 2 minutes in a water bath at 95°C.  Allow
      the sample to cool and add 50 mL distilled water, 15 mL of KMnO/
      solution (6.5)  and 8 mL of potassium persulfate solution (6.6) to
      each bottle and return to the water bath for 30 minutes.   Cool and
      add 6 mL of sodium chloride-hydroxylamine sulfate solution (6.4) to
      reduce the excess  permanganate.  Add 50 mL of distilled water.
      Treating each bottle individually, add 5 mL of stannous sulfate
      solution (6.3) and immediately attach the bottle to the aeration
      apparatus.  At this point the sample is allowed to stand quietly
      without manual agitacion.  The circulating pump, which has previously
      been adjusted to a rate of 1 liter per minute, is allowed to run
      continuously.  The absorbance, as exhibited either on the
      spectrophotometer or the recorder, will increase and reach maximum

                                   D-66                                ILM01.0

-------
      within 30 seconds.  As soon as the recorder pen levels off,
      approximately 1 minute, open the bypass valve and continue the
      aeration until the absorbance returns to its minimum value (see Note
      4).  Close the bypass valve, remove the fritted tubing from the BOD
      bottle and continue the aeration.  Proceed with the standards and
      construct a standard curve by plotting peak height versus micrograms
      of mercury

      NOTE 4:     Because of the toxic nature of mercury vapor, precaution
      must be taken to avoid its inhalation.  Therefore, a bypass has been
      included in the system to either vent the mercury vapor into an
      exhaust hood or pass the vapor through some absorbing media, such as:
      a) equal volumes of 0.1 N KMnO^ and 10% i^SO^, or b) 0.25% iodine in
      a 3% KI solution.  A specially treated charcoal that will absorb
      mercury vapor is also available.

8.     Procedure

8.1   Weigh a representative 0.2 g portion of wet sample and place in the
      bottom of a BOD bottle.  Add 5 mL of sulfuric acid (6.1)  and 2.5 mL
      of concentrated nitric acid (6.2) mixing after each addition.' Heat
      two minutes in a water bath at 95°C.   Cool,  add 50 mL distilled
      water,  15 mL potassium permanganate solution (6.5) and 8  mL of
      potassium persulfate solution (6.6) to each sample bottle.   Mix
      thoroughly and place in the water bath for 30 minutes at  95°C.  Cool
      and add 6 mL of sodium chloride-hydroxylamine sulfate (6.4) to reduce
      the excess permanganate.   Add 55 mL of distilled water.   Treating
      each bottle individually,  purge the head space of the sample bottle
      for at least one minute and add 5 mL of stannous sulfate  (6.3) and
      immediately attach the bottle to the aeration apparatus.   Continue as
      described under (7.1)

8.2   An alternate digestion procedure employing an autoclave may also be
      used.  In this method 5 mL of cone.  HoSO^ and 2 mL of cone.  HNO^
      are added to the 0.2 g of sample.  5 mL of saturated KMnO^ solution
      and 8 mL of potassium persulfate solution =>re added and the bottle is
      covered with a piece of aluminum foil.  The sample is autoclaved at
      121°C and 15 Ibs.  for 15 minutes.   Cool,  make up to a volume of 100
      mL with distilled water and add 6 mL of sodium chloride-hydroxylamine
      sulfate solution (6.4) to reduce the excess permanganate.  Purge the
      head space of the sample bottle for at least one minute and continue
      as described under (7.1)

9.     Calculations

9.1   Measure the peak height of the unknown from the chart and read the
      mercury value from the standard curve

9.2   Calculate the mercury concentration in the sample by the  formula:

                    ug Hg in the aliquot
          US  g/g = wt of the aliquot in gms
                    (based upon dry wt of the sample)
                                                                       ILMOL.O

-------
9.3   Report mercury concentrations as described for aqueous mercury
      samples converted to units of rag/kg.  The sample result or the
      detection limit for each sample must be corrected for sample weight
      and % solids before reporting.

Bibliography

1.   Bishop, J.   N.,  "Mercury in Sediments",  Ontario Water Resources Comm.,
     Toronto, Ontario,  Canada,  1971

2.   Salraa,  M.,  private communication,  EPA Cal/Nev.   Basin Office,  Almeda,
     California

3.   "Interim Methods for the Sampling and Analysis  of Priority Pollutants
     in Sediments and Fish Tissue,"  USEPA  Environmental Monitoring and
     Support Laboratory,  Cincinnati,  Ohio,  August  1977,  Revised October 1980

4.   Op.   cit.   (#3),  Methods 245.1  or 245.2
                                  D-68                                ILM01.0

-------
                   PART E  - METHODS FOR CYANIDE ANALYSIS
Method                                                            Page No.

Method for Total Cyanide Analysis in Water*
Method 335.2 CLP-M                                                 D-70

Method for Total Cyanide Analysis in Soil/Sediment*
Method 335.2 CLP-M                                                 D-79

Method for Total Cyanide Analysis by Midi Distillation
Method 335.2 CLP-M                                                 D-91
 A bibliography citing method references follows the method.
&
 CLP-M Modified for the Contract Laboratory Program.

                                   D-69                                ILM01.0

-------
                 METHOD FOR TOTAL CYANIDE ANALYSIS IN WATER

                          CYANIDE, TOTAL  (in Water)
Method  335.2 CLP-M   (Titrimetric; Manual  Spectrophotometric; Semi-Automated
                             Spectrophotometrie)

1.    Scope and Application

1.1   This method is applicable to the determination of cyanide in
      drinking, surface and saline waters, domestic and industrial wastes.

1.2   The titration procedure using silver nitrate with p-
      dimethylaminobenzalrhodanine indicator is used for measuring
      concentrations of cyanide exceeding 1 mg/L (0.25 mg/250 mL of
      absorbing liquid).  (Option A, 8.2).

1.3   The manual colorometric procedure is used for concentrations  below 1
      mg/L of cyanide and is sensitive to about 0.01 mg/L.  (Option B,
      8.3).

1.4   The working range of the semi-automated Spectrophotometric method is
      0.020 to 0.200 mg/L.  Higher level samples must be diluted to fall
      within the working range.  (Option C, 8.4).

2.    Summary of Method

2.1   The cyanide as (HRN) hydrocyanic acid (HCN) is released from cyanide
      complexes by means of a reflux-distillation operation and absorbed In
      a scrubber containing sodium hydroxide solution.  The cyanide ion in
      the absorbing solution is then determined by volumetric titration or
      colorimetrically.

2.2   In the colorimetric measurement the cyanide is converted to cyanogen
      chloride, CNC1, by reaction with chloramine-T at a pH less than 8
      without hydrolyzing to the cyanate.  After the reaction is complete,
      color is formed on the addition of pyridine-pyrazolone or
      pyridinebarbituric acid reagent.  The absorbance is read at 620 nm
      when using pyridine-pyrazolone or 578 nm for pyridine-barbituric
      acid.  To obtain colors of comparable intensity, it is essential to
      have the same salt content in both the sample and the standards.

2.3   The titimetric measurement uses a standard solution of silver nitrate
      to titrate cyanide in the presence of a silver sensitive indicator.

3.    Definitions

3.1   Cyanide  is defined as cyanide ion and complex cyanides converted to
      hydrocyanic acid  (HCN) by reaction in a reflux system of a mineral
      acid in  the presence of magnesium ion.
 CLP-M Modified  for the Contract Laboratory Program.

                                   D-70                                ILM01.0

-------
 4.     Sample Handling and Preservation

 4.1    All  bottles  must be thoroughly cleansed and rinsed  to  remove  soluble
       material  from containers.

 4.2    Oxidizing agents such as chlorine decompose most of the cyanides.
       Test a drop  of the  sample with potassium iodide-starch test paper
       (Kl-starch paper);  a. blue color indicates the need  for treatment.
       Add  ascorbic acid,  a few crystals at a time, until  a drop of  sample
       produces  no  color on the indicator paper.  Then add amadditional 0.6
       g of ascorbic acid  for each liter of sample volume.

 4.3    Samples are  preserved with 2 mL of 10 N sodium hydroxide per  liter of
       sample  (pH>  12)  at  the time of collection (Exhibit  D,   Section II).

 4.4    Samples must be  stored at 4°C(±2°C) and must be analyzed within the
       holding time specified in Exhibit D,  Section II.

 5.     Interferences

 5.1    Interferences are eliminated or reduced by using the distillation
       procedure  described  in Procedure 8.1.

 5.2-    Sulfides adversely affect the colorimetric and titration procedures.
       If a  drop  of the  distillate on lead acetate test paper indicates the
       presence of  sulfides, treat 25 mL more of the sample than that
       required for the  cyanide determination with powdered cadmium
       carbonate.   Yellow cadmium sulfide precipitates if  the sample
       contains sulfide.  Repeat this operation until a drop of the treated
       sample solution does not darken the lead acetate test paper.  Filter
       the  solution through a dry filter paper into a dry beaker, and from
       the  filtrate measure the sample to be used for analysis.   Avoid a
       large excess  of cadmium carbonate and a long contact time in order to
       minimize a loss by complexation or occlusion of cyanide on the
       precipitated material.   Sulfides should be removed prior to
       preservation with sodium hydroxide as described in 4.3.

 5.3    The presence  of surfactants may cause the sample to foam during
       refluxing.   If this occurs,  the addition of an agent such as Dow
       Corning 544  antifoam agent will prevent the foam from collecting in
       the condenser.  Fatty acids will distill and form soaps under
       alkaline titration conditions,  making the end point almost impossible
       to detect.   When this occurs,  one of the spectrophotometric methods
       should be used.

6.    Apparatus

6.1   Reflux distillation apparatus  such as shown in Figure  1 or Figure 2.
      The  boiling  flask should be of 1 liter size with inlet tube and
      provision for condenser.   The  gas absorber may be a Fisher-Milligan
      scrubber.

6.2   Microburet, 5.0 mL (for  titration)


                                  D-71                               ILM01.0

-------
6.3    Spectrophotometer suitable for measurements at 578 run or  620 nra with
       a  1.0 cm cell or larger (for manual spectrophotometric method).

6.4    Technicon AA II System or equivalent instrumentation, (for automated
       spectrophotometric method) including:

       6.4.1   Sampler

       6.4.2   Pump III

       6.4.3   Cyanide Manifold  (Figure 3)

       6.4.4   SCIC Colorimeter with 15 mm flowcells and 570 nm  filters

       6.4.5   Recorder

       6.4.6   Data System (optional)

       6.4."7   Glass or plastic tubes for the sampler

7.    Reagents

7.1   Distillation and Preparation Reagents

      7.1.1   Sodium hydroxide solution, 1.25N:  Dissolve  50 g  of NaOH  in
              distilled water, and dilute to 1 liter with  distilled water.

      7.1.2   Cadmium carbonate:  powdered

       7.1.3   Ascorbic acid:  crystals

       7.1.4   Sulfuric acid:  concentrated

       7.1.5   Magnesium chloride solution:  Weight 510 g of MgC^'Sl^O  into
              a 1000 mL flask, dissolved and dilute to 1 liter  with
              distilled water.

7.2   Stock Standards  and Titration Reagents

      7.2.1   Stock cyanide solution:  Dissolve 2.51 g of  KCN and 2 g KOH
              in 1 liter of distilled water.   Standardize  with  0.0192 N
              AgN03.

      7.2.2   Standard cyanide solution, intermediate:  Dilute  50.0 mL of
              stock (1 mL = 1 mg CN) to 1000 mL with distilled  water.

      7.2.3   Standard cyanide solution:  Prepare fresh daily by diluting
              100.0 mL of intermediate cyanide solution to 1000 mL with
              distilled water and store in a glass stoppered bottle.  1 mL
              = 5.0 ug CN (5.0 mg/L).

      7.2.4   Standard silver nitrate solution, 0.0192 N:   Prepare by
              crushing approximately 5 g AgN03 crystals and drying to
              constant weight at 40°C.  Weight out 3.2647  g of  dried AgN03,

                                   D-72                                ILM01.0

-------
              dissolve in distilled water, and dilute to 1000 mL (1 mL - 1
              rag CN).

      7.2.5   Rhodanine indicator:  Dissolve 20 mg of p-dimethyl-
              aminobenzalrhodanine in 100 mL of acetone.

      7.2.6   Sodium hydroxide solution, 0.25 N:  Dissolve 10 g or NaOH in
              distilled water and dilute to 1 liter.

7.3 Manual  Spectrophotometric Reagents

      7.3.1   Sodium dihydrogenphosphate, 1 M:  Dissolve 138 g of
              Nat^PO^'H^O in a liter of distilled water.  Refrigerate this
              solution.

      7.3.2   Chloramine-T solution:  Dissolve 1.0 g of white, water
              soluble chloramine-T in 100 mL of distilled water and
              refrigerate until ready to use.  Prepare fresh weekly.

      7.3.3   Color Reagent-One of the following may be used:

              7.3.3.1    Pyridine-barbituric acid reagent:   Place 15 g of
                         barbituric acid in a 250 mL volumetric flask and
                         add just enough distilled water to wash the sides
                         of the flask and wet the barbituric acid.   Add 75
                         mL of pyridine  and mix.   Add 15 mL of HC1  (sp gr
                         1.19),  mix,  and cool to room temperature.   Dilute
                         to 250 mL with  distilled water and mix.  This
                         reagent is stable for approximately six months if
                         stored in a cool,  dark place.

              7.3.3.2    Pyridine-pyrazolone solution:?.3.3.2.1 3-Methyl-l-
                         phenyl-2-pyrazolin-5-one reagent,  saturated
                         solution:   Add  0.25 g of 3-methyl-l-phenyl-2-
                         pyrazolin-5-one to  50 mL of distilled water,  heat
                         to  60°C with stirring.   Cool to room temperature.

                        7.3.3.2.1     3-Methyl-lphenyl-2-pyrazolin-5-one
                                      reagent, saturated  solution:  Add
                                      0.25 g of 3-methyl-l-phenyl-2-
                                      pyrazolin-5-one to  50 mL of distilled
                                      water, heat  to  60°C with stirring.
                                      Cool to  room temperature.

                        7.3.3.2.2     3,3'Dimethyl-l,l'-diphenyl  [4,4'-bi-2
                                      pyrazolin]-5,5'dione  (bispyrazolone):
                                      Dissolve 0.01 g of bispyrazolone in
                                      10 mL of pyridine.

                        7.3.3.2.3     Pour solution  (7.3.3.2.1) through
                                      nonacid-washed  filter paper.  Collect
                                      the filtrate.   Through  the same
                                      filter paper pour solution
                                      (7.3.3.2.2)  collecting  the filtrate

                                  D-73                                ILM01.0

-------
                                      in the same container as filtrate
                                      from (7.3.3.2.1).   Mix until the
                                      filtrates are homogeneous.  The mixed
                                      reagent develops a pink color but
                                      this does not affect the color
                                      production with cyanide if used
                                      within 24 hours of preparation.

7.4   Semi -Automated Spectropho tome trie Reagents

      7.4.1   Chloramine-T solution:  Dissolve 0.40 g of chloramine-T  in
              distilled water and dilute to 100 mL.  Prepare  fresh daily.
      7.4.2   Phosphate buffer:  Dissolve 138 g of Nal^PO^'l^O  in distilled
              water and dilute to 1 liter.  Add 0.5 mL of Brij-35
              (available from Technicon) .  Store at 4°C(±2°C).

      7.4.3   Pyridine -barbituric acid solution:  Transfer 15 g of
              barbituric acid into a 1 liter volumetric flask.  Add about
              100 mL of distilled water  and swirl the flask.  Add 74 mL of
              pyridine and mix.  Add 15  mL of concentrated HC1  and 'mix.
              Dilute to about 900 mL with distilled water and mix until the
              barbituric acid is dissolved.  Dilute to 1 liter  with
              distilled water.  Store at 4°C(±2°C) .

      7.4.4   Sampler wash:  Dissolve 10 g of NaOH in distilled water  and
              dilute to 1 liter.

8 .     Procedure

8.1   Distillation

      8.1.1   Place 500 mL of sample in  the 1 liter boiling flask.  Add 50
              mL, of sodium hydroxide (7.1.1) to the absorbing  tube and
              dilute if necessary with distilled water to obtain  an
              adequate depth of liquid in the absorber.  Connect  the
              boiling flask, condenser,  absorber and trap in the  train.

          8.1.2    Start a slow stream of air  entering the boiling flask by
                  adjusting the vacuum source.   Adjust the  vacuum so that
                  approximately one bubble of air per second enters the
                  boiling flask through the air inlet tube.

                  NOTE:   The bubble rate will not remain constant after the
                  reagents  have been added and while  heat is being applied
                  to the  flask.   It will be necessary to readjust the  air
                  rate  occasionally to prevent the solution in the boiling
                  flask from backing up into  the air  inlet  tube.

      8.1.3   Slowly add 25 mL concentrated sulfuric acid (7.1.4)  through
              the air inlet tube.  Rinse the tube with distilled  water and
              allow the airflow to mix the flask contents for 3 minutes.
              Pour 20 mL of magnesium chloride solution (7.1.5) into the
              air inlet and wash down with a stream of water. .

                                   D-74                                ILM01.0

-------
      8.1.4    Heat the solution to boiling,  taking care to prevent the
               solution from backing up into  and overflowing from the air
               inlet tube.   Reflux for one  hour.   Turn off heat and continue
               the airflow for at least 15  minutes.   After cooling the
               boiling flask,  disconnect absorber and close off the vacuum
               source.

      8.1.5    Drain the solution from the  absorber  into a 250  mL volumetric
               flask and bring up to volume with distilled water washings
               from the absorber tube.

8.2   Titrimetric Determination  (Option A)

      8.2.1    If the sample  contains more  than 1 mg of CN,  transfer the
               distillate,  or  a suitable aliquot diluted to 250 mL,  to a  500
               mL Erlenmeyer  flask.   Add 10-12  drops of the benzalrhodanine
               indicator.

      8.2.2    Titrate  with standard silver nitrate  to the first change in
               color from yellow to  brownish-pink.   Titrate a distilled
               water blank  using the same amount  of  sodium hydroxide and
               indicator as in the sample.

      8.2.3    The analyst  should familiarize himself with the  end point:  of
               the titration  and the amount of  indicator to be  used before
               actually titrating the samples.   A 5  or 10 mL microburet may
               be conveniently used to obtain a more precise titration.

8.3   Manual Spectrophotometric Determination  (Option B)

      8.3.1    Withdraw 50  mL  or less of the solution from the  flask and
               transfer to  a  100 mL volumetric  flask.   If less  than 50 mL is
               taken,  dilute  to 50 mL with  0.25 N sodium hydroxide solution
               (7.2.6).   Add  15.0 mL of sodium  phosphate solution (7.3.1)
               and mix.   The dilution factor must be reported on Form XIV.

               8.3.1.1    Pyridine-barbituric acid method:  Add  2 mL of
                         chloramine-T  (7.3.2)  and mix.  After 1  to 2
                         minutes, add 5 mL of  pyridine-barbituric acid
                         solution (7.3.3.1) and mix.  Dilute to mark with
                         distilled water and mix again.  Allow  8 minutes
                         for color development then read absorbance at 578
                         nm in a 1 cm cell within 15 minutes.

               8.3.1.2    Pyridine-pyrazolone method:  Add 0.5 mL of
                         chloramine-T (7.3.2)  and mix.  After I to 2
                         minutes, add 5 mL of pyridine-pyrazolone solution
                         (7.3.3.2) and mix.  Dilute to mark with distilled
                         water and mix again.  After 40 minutes, read
                         absorbance at 620 nm  in a 1 cm cell.  NOTE:  More
                         than 0.5 mL of chloramine-T will prevent the color
                         from developing wich pyridine-pyrazolone.
                                   D-75                               ILM01.0

-------
       8.3.2    Prepare a minimum of 3 standards  and a blank by pipetting
               suitable volumes  of standard solution into  250  mL volumetric
               flasks.   NOTE:  One calibration standard must be at the
               Contract Required Detection  Limit (CRDL).   To each standard,
               add 50 mL of 1.25 N sodium hydroxide and dilute to 250 mL
               with distilled water.   Standards  must bracket the
               concentration of  the samples.  If dilution  is required,  use
               the blank solution.

               As  an example, standard solutions  could be  prepared as
               follows:

                        mL of Standard Solution      Cone, ug CN
                            (1.0 - 5 ug CN)          per 250 mL

                                  0                     Blank
                                  1.0                      5
                                  2.0                     10
                                  5.0                     25
                                 10.0                     50
                                 15.0                     75       '
                                 20.0                    100

               8.3.2.1    It is not imperative that all standards be
                         distilled in the same manner as the samples.  At
                         least one standard (mid-range) must be distilled
                         and compared to similar values on the curve to
                         ensure  that the distillation technique is
                         reliable.  If the distilled standard does not
                         agree within +15% of the undistilled standards,
                         the operator should find and correct the cause of
                         the apparent error before proceeding.

               8.3.2.2    Prepare a standard curve by plotting absorbance of
                         standard vs.  cyanide  concentrations (per 250 mL).

8.4   Semi-Automated Spectrophotometric  Determination (Option  C)

      8.4.1    Set up the manifold as shown in Figure 3.  Pump the reagents
               through the system until a steady baseline is obtained.

      8.4.2    Calibration standards:  Prepare a blank and at least three
               calibration standards over the range of the analysis.  One
               calibration standard must be at the CRDL.  For a working
               range of 0-200 ug/L, the following standards may be used:
                                  D-76                                ILM01.0

-------
                          mL Standard Solution        Concentration
                        (7.2.3") diluted to 1 liter       ug CN/L

                                      0                     0
                                    4.0                    20
                                   10.0                    50
                                   20.0                   100
                                   30.0                   150
                                   40.0                   200

              Add 10 g of NaOH  to each standard.  Store at 4°C(±2°C)

      8.4.3   Place calibration standards, blanks, and control standards in
              the sampler tray, followed by distilled samples, distilled
              duplicates, distilled standards, distilled spikes, and
              distilled blanks.

      8.4.4   When a steady reagent baseline is obtained and before
              starting the sampler, adjust the baseline u^-'.ng the
              appropriate knob  on the colorimeter.  Aspirate a. calibration
              standard and adjust the STD CAL dial on the colorimeter until
              the desired signal is obtained.  Record the STD CAL value.
              Re-establish the  baseline and proceed to analyze calibration
              standards, blanks, control standards, distilled samples,  and
              distilled QC audits.

9.     Calculations

9.1   Using the titrimetric procedure,  calculate concentration of CN as
      follows:

                        (A-B)  1.000 mL/L     	250 mL	
          CN,  rag/L -     mL orig.  sample  x  mL of aliquot titrated

          WHERE:  A - volume of AgNO-j for titration of sample
                     (1 mL - 1 mg Ag)
                  B = volume of AgNO-j for titration of blank
                     (1 mL - 1 mg Ag)

          AND:    250 mL = distillate volume (See 8.1.5)
                  1000 mL = conversion mL to L
                  mL original sample (See 8.1.1)
                  mL of aliquot titrated  (See 8.2.1)

9.2   If the semi-automated method is used, measure the peak heights of the
      calibration standards (visually or using a data system) and calculate
      a linear regression equation.   Apply the equation to the samples and
      QC audits to determine the cyanide concentration in the distillates.
      To determine the concentration of cyanide in the original sample,
      MULTIPLY THE RESULTS BY ONE-HALF (since the original volume was  500
      mL and the distillate volume was 250 mL).   Also,  correct for,  and
      report on Form XIV,  any dilutions which were made before or after
      distillation.


                                   D-77                               ILM01.0

-------
      The minimum concentration that can be reported from the calibration
      curve is 20 ug/L that corresponds to 10 ug/L in a sample that has
      been distilled.

9.3   If the colorimetric procedure is used, calculate the cyanide, in
      ug/L, in the original sample as follows:

                      A x 1.000 mL/L    x   50 mL
          CN,  ug/L        B                 C
                                                            i
          WHERE:   A - ug CN read from standard curve (per 250 mL)
                  B - mL of original sample for distillation (See 8.1.1)
                  C - mL taken for colorimetric analysis (See 8.3.1)

          AND:     50 mL - volume of original sample aliquot (See 8.3.1)
                  1000 mL/L - conversion mL to L

          The  minimum value that can be substituted for A is 5 ug per 250
          mL.   That yields i concentration of 10 ug/L in the distilled
          sample.

Bibliography

1.    Methods  for "Chemical Analysis of Water and Wastes",  March 1979, EPA
      publication #600/4-79-02.

2.    "Operation RN Manual for Technicon Auto Analyzer IIC System", 1980.
      Technical publication #TA9-0460-00.    Technicon Industrial Systems,
      Tarrytown,  NY, 10591.

3.    "Users Guide for the Continuous Flow Analyzer Automation System",
      EMSL U.S. EPA, Cincinnati,  OH (1981).

4.    Interim Methods for the Sampling and Analysis of Priority Pollutants
      in Sediments and Fish Tissue," USEPA Environmental Monitoring and
      Support Laboratory, Cincinnati,  Ohio, August 1977, Revised October
      1980.

5.    Op.  cit.  (#4), Methods 335.2.
                                   D-78                                 ILM01.0

-------
              METHOD FOR TOTAL CYANIDE ANALYSIS IN SOIL/SEDIMENT
                        CYANIDE, TOTAL (in Sediments)

                        P-M* (Titrimetric; Manual Spec
                      Semi-Automated Spec tropho tome tric)
Method 335.2 CLP-M* (Titrimetric; Manual Spectrophotometric;
 1.     Scope and Application

 1.1   This method is applicable to the  determination of cyanide in
       sediments and other solids.

 1.2   The  detection limit is  dependent  upon  the weight  of sample taken for
       analys is.

 2.     Summary of Method

 2.1   The  cyanide as hydrocyanic acid (HCN)  is released from cyanide
       complexes  by means  of a reflux-distillation operation and absorbed in
       a scrubber containing sodium hydroxide solution.   The cyanide-ion in
       the  absorbing solution  is then determined by volumetric titration or
       colorimetrically.

 2.2    In the colorimetric measurement the cyanide is converted  to  cyanogen
       chloride,  CNC1, by  reaction with chloramine-T at  a pH less than  8
       without hydrolyzing to  the cyanate.  After the reaction is complete,
       color is formed on  the  addition of pyridine-pyrazolone or pyridine-
       barbituric  acid reagent.  The absorbance is read  at 620 nm when  using
       pyridine-pyrazolone for 578 nm for pyridine-barbituric acid.  To
       obtain colors of  comparable intensity, it is essential to have the
       same salt content in  both the sample and the standards.

 2.3    The  titrimetric measurement uses a standard solution of silver
       nitrate to  titrate cyanide in the presence of a silver sensitive
       indicator.

 3.     Definitions

 3.1    Cyanide is  defined as cyanide ion and complex cyanides converted to
      hydrocyanic acid  (HCN) by reaction in a reflux system of a mineral
       acid in the presence of magnesium ion.

4.     Sample Handlins and Preservation

4.1   Samples must be stored at 4°C(+2°C) and must be analyzed within the
      holding time specified in Exhibit D,  Section II.

4.2   Samples are not dried prior to analysis.   A separate percent solids
      determination must be made in accordance with the procedure in Part
      F.
 CLP-M  Modified for the Contract Laboratory Program.

                                   D-79                                ILM01.0

-------
5.    Interferences

5.1   Interferences are eliminated or reduced by using the distillation
      procedure described in Procedure 8.1.

5.2   Sulfides adversely affect the colorimetric and titration procedures.

5.3   The presence of surfactants may cause the sample to foam during
      refluxing.  If this occurs, the addition of an agent such as DOW
      Corning 544 antifoam agent will prevent the foam from collecting in
      the condenser.  Fatty acids will distill and form soaps under the
      alkaline titration conditions, making the end point almost: impossible
      to detect.  When this occurs,  one of the spectrophotometric methods
      should be used.

6.    Apparatus

6.1   Reflux distillation apparatus such as shown in Figure 1 or Figure 2.
      The boiling flask should be of 1 liter size with inlet tube and
      provision for condenser.   The gas absorber may be a Fisher-Milligan
      scrubber.

6.2   Microburet,  5.0 mL (for titration)

6.3   Spectrophotometer suitable for measurements at 578 nm or 620 nm with
      a 1.0 cm cell or larger.

6.4   Technicon AA II System or equivalent instrumentation (for automated
      spectrophotometric method) including:

      6.4.1    Sampler

      6.4.2    Pump  III

      6.4.3    Cyanide Manifold  (Figure  3)

      6.4.4    SCIC  Colorimeter  with 15  mm  flowcells and 570 nm  filters

      6.4.5    Recorder

      6.4.6    Data  System  (optional)

      6.4.7    Glass or plastic  tubes  for the  sampler

7.    Reagents

7.1   Distillation and Preparation Reagents

      7.1.1    Sodium hydroxide  solution,  1.25N:   Dissolve  50  g  of  NaOH  in
               distilled water,  and  dilute  to  1  liter with  distilled water.

      7.1.2    Cadmium  carbonate:  powdered

      7.1.3    Ascorbic acid:   crystals

                                   D-80                                ILM01.0

-------
      7.1.4    Sulfuric  acid:   concentrated

      7.1.5    Magnesium chloride  solution:  Weigh  510  g  of MgClZ.ei^O  into
               a  1000  mL flask,  dissolve  and dilute to  1  liter with
               distilled water.

7.2   Stock Standards and Titration Reagents

      7.2.1    Stock cyanide  solution:  Dissolve 2.51 g of KCN and 2 g  KOH
               in 1  liter of  distilled water.  Standardize with 0.0192  N
               AgN03.

      7.2.2    Standard  cyanide solution,  intermediate:   Dilute 50.0 mL of
               stock (1  ml -  1  mg  CN) to  1000 mL with distilled water (1 mL
               -  50.0  ug).

      7.2.3    Standard  cyanide solution:  Prepare  fresh  daily by diluting
               100.0 mL  of intermediate cyanide solution  to 1000 mL with
               distilled water  and store  in a glass stoppered bottle.   1 mL
               -  5.0 ug  CN (5.0 mg/L).

      7.2.4    Standard  silver  nitrate solution, 0.0192 N:  Prepare by
               crushing  approximately 5 g AgNOo crystals  and drying to
               constant  weight  at  40°C.   Weigh out  3.2647 g of dried
               dissolve  in distilled water, and dilute  to 1000 mL  (1 mL = 1
               mg CN).

      7.2.5    Rhodanine indicator:  Dissolve 20 mg of  p-dimethy1-amino-
               benzalrhodanine  in  100 mL  acetone.

7.3   Manual Spectrophotometric Reagents

      7.3.1    Sodium  dihydrogenphosphate, 1 M:  Dissolve 138 g of
               NaH2P04'H20 in 1 liter of  distilled  water.  Refrigerate  this
               solution.

      7.3.2    Chloramine-T solution:  Dissolve 1.0 g of  white, water
               soluble Chloramine-T in 100 mL of distilled water and
               refrigerate until ready to use.  Prepare fresh weekly.

      7.3.3    Color reagent  -  One of the following may be used:

               7.3.3.1   Pyridine-barbituric acid  reagent:   Place 15 g of
                         barbituric acid  in a 250 mL volumetric flask and
                         add just enough  distilled water to wash the sides
                         of  the flask and wet the barbituric acid.  Add 75
                         mL of pyridine  and mix.   Add  15 mL of HC1 (sp gr
                         1.19), mix, and  cool to room  temperature.  Dilute
                         to  250 mL with  distilled water and mix.   This
                         reagent  is stable for approximately six months if
                         stored in a cool, dark place.

               7.3.3.2    Pyridine-pyrazolone solution:
                                   D-81                                ILM01.0

-------
                         7.3.3.2.1    3-Methyl-1-phenyl- 2-pyrazolin-5-one
                                      reagent, saturated solution:  Add
                                      0.25 g of 3-methyl-l-phenyl-2-
                                      pyrazolin-5-one to 50 mL of distilled
                                      water, heat to 60°C with stirring.
                                      Cool to room temperature.

                         7.3.3.2.2    3,3'Dimethyl-1,1'-diphenyl-[4,4'-bi-
                                      2-pyrazolin]-5,5'dione
                                      (bispyrazolone):  Dissolve 0.01 g of
                                      bispyrazolone in 10 mL of pyridine.

                         7.3.3.2.3    Pour solution (7.3.3.2.1) through
                                      non-acid-washed filter paper.
                                      Collect the filtrate.  Through the
                                      same filter paper pour solution
                                      (7.3.3.2.2) collecting the filtrate
                                      in the same container as filtrate
                                      from (7.3.3.2.1).   Mix until the
                                      filtrates are homogeneous.   The mixed
                                      reagent develops a pink color" but
                                      this does not affect  the color
                                      production with cyanide if used
                                      within 24 hours of preparation.

7.4   Semi-Automated Spectrophotometric Reagents

      7.4.1   Chloramine-T solution:  Dissolve 0.40 g of chloramine-T in
              distilled water and dilute to 100 mL.  Prepare fresh daily.

      7.4.2   Phosphate Buffer:  Dissolve 138 g of NaH2P04'H2O in distilled
              water and dilute to 1 liter.  Add 0.5 mL of Brij-35
              (available from Technicon).  Store at 4°C.

      7.4.3   Pyridine-barbituric acid solution:  Transfer 15 g of
              barbituric acid into a 1 liter volumetric flask.  Add about
              100 mL of distilled water and swirl the flask.  Add 74 mL of
              pyridine and mix.  Add 15 mL of cone.  HC1 mix until the
              barbituric acid is dissolved.  Dilute to 1 liter with
              distilled water.  Store at 4°C.

      7.4.4   Sampler Wash:  Dissolve 10 g of NaOH in distilled water and
              dilute to 1 liter.

8.     Procedure

8.1   Distillation

      8.1.1   Accurately weigh a representative 1-5 g portion of wet sample
              and transfer it to a boiling flask.  Add 500 mL of distilled
              water.  Shake or stir the sample so that it is dispersed.

      8.1.2   Add 50 mL of sodium hydroxide  (7.1.1) to the absorbing tube
              and dilute if necessary with distilled water to obtain an

                                   D-82                               ILM01.0

-------
               adequate depth of liquid in the absorber.  Connect the
               boiling flask, condenser, absorber, and trap in the train.

       8.1.3   Start a slow stream of air entering the boiling flask by
               adjusting the vacuum source.   Adjust the vacuum so that
               approximately one bubble of air per second enters the boiling
               flask through the air inlet tube.

                   NOTE:   The bubble  rate will  not remain constant  after the
                   reagents  have been added and while heat is. being applied
                   to  the  flask.   It  will be necessary  to readjust  the air
                   rate  occasionally  to  prevent the solution in the boiling
                   flask from backing up into the  air inlet tube.

       8.1.4   Slowly add  25 mL of cone, sulfuric  acid (7.1.4) through the
               air inlet tube.  Rinse the  tube  with distilled water and
               allow the airflow to mix the  flask  contents for 3 minutes.
               Pour 20 mL  of magnesium .chloride solution (7.1.5)  into the
               air inlet and wash down with  a stream of water.

       8.1.5   Heat the solution to boiling,  taking care to prevent'the
               solution from backing  up  and  overflowing into the air inlet
               tube.   Reflux for  one  hour.  Turn off  heat and continue the
               airflow for at least 15 minutes.  After  cooling the  boiling
               flask,  disconnect  absorber and close off the vacuum source.

       8.1.6   Drain the solution  from the absorber into a 250 raL volumetric
               flask and bring up  to  volume with distilled water  washings
               from the  absorber tube.

8.2    Titrimetric Determination (Option A)

       8.2.1    If  the  sample contains more than 1  mg of CN,  transfer  the
               distillate, or a suitable aliquot diluted to 250 mL,  to  a 500
               mL  Erlenmeyer flask.   Add 10-12  drops of the benzalrhodanine
               indicator.

       8.2.2    Titrate with  standard  silver nitrate to  the first  change in
               color from yellow to brownish-pink.  Titrate a  distilled
               water blank using the  same amount of sodium hydroxide  and
               indicator as  in the  sample.

       8.2.3    The  analyst should  familiarize himself with the end  point of
               the  titration and the  amount of  indicator  to be used before
               actually  titrating  the samples.  A  5 or  10 mL microburet may
               be conveniently used to obtain a more precise titration.

8.3   Manual Spectrophotometric Determination (Option B)

      8.3.1    Withdraw  50 mL or less of the solution from the flask  and
               transfer  to a 100 mL volumetric  flask.   If less  than 50 mL  is
               taken, dilute to 50 mL with 0.25 N  sodium  hydroxide  solution
               (7.2.6).  Add 15.0  mL  of  sodium phosphate  solution (7.3.2)
               and mix.

                                   D-83                                 ILM01.0

-------
         8.3.1.1    Pyridlne-barbituric acid method:  Add 2 mL of
                   Chloramine-T (7.3.2) and mix.  After 1 to 2
                   minutes, add 5 mL of pyridine-barbituric acid
                   solution (7.3.3.1) and mix.  Dilute to mark with
                   distilled water and mix again.  Allow 8 minutes
                   for color development then read absorbance at 578
                   nm in a 1 cm cell within 15 minutes.

         8.3.1.2    Pyridine-pyrazolone method:  Add 0.5 mL of
                   chloramine-T (7.3.2) and mix.  After 1 to 2
                   minutes add 5 mL of pyridine-pyrazolone solution
                   (7.3.3.2) and mix.  Dilute to mark with distilled
                   water and mix again.  After 40 minutes read
                   absorbance at 620 nm in a 1 cm cell.

                   NOTE:  More than 0.5 mL of chloramine-T will
                   prevent the color from developing with pyridine-
                   pyrazolone.

8.3.2    Prepare  a minimum of three standards and a blank by pipetting
         suitable volumes of standard solution into 250 mL volumetric
         flasks.

    NOTE:  One calibration standard,  must be made at the CRDL.   To
    each standard add 50 mL of 1.25 N sodium hydroxide and dilute to
    250 mL with distilled water.   Standards must bracket the
    concentrations of the sample.   If dilution is required,  use the
    blank solution.

    As an example,  standard solutions could be prepared as follows:

                  mL of Standard Solution       Cone,  ug CN
                      (1.0 - 5 ug CN")            per 250 mL

                            0                      Blank
                            1.0                       5
                            2.0                      10
                            5.0                      25
                           10.0                      50
                           15.0                      75
                           20.0                     100

         8.3.2.1    It is not imperative that all standards be
                   distilled in the same manner as the samples.  At
                   least one standard (mid-range) must be distilled
                   and compared to similar values on the curve to
                   insure that the distillation technique is
                   reliable.  If the distilled standard does not
                   agree within +15% of the undistilled standards the
                   operator should find and correct the cause of the
                   apparent error before proceeding.

         8.3.2.2    Prepare a standard curve by plotting absorbance of
                   standard vs. cyanide concentrations (per 250 mL)

                             D-84                                ILM01.0

-------
8.4   Serai-Automated Spectrophotometric Determination (Option C)

      8.4.1    Set up the manifold as shown  in Figure 3.  Pump the reagents
               through  the system until a steady baseline is obtained.

      8.4.2    Calibration standards:  Prepare a blank and at least three
               calibration standards over the range of the analysis.  One
               calibration standard must be  at the CRDL.  For a working
               range of 0-200 ug/L, the following standards may be used:

                          mL Standard Solution           Concentration
                        (1.2. 3") diluted to 1 liter           ug CN/L

                                   0                            0
                                   4.0                         20
                                   10.0                        50
                                   20.0                       10O
                                   30.0                       150
                                   40.0                       200

                  Add 10 g of NaOH to each standard.   Store at 4°C(±2°C).

          8.4.3   Place calibration standards,  blanks,  and control
                  standards in the sampler tray,  followed by distilled
                  samples,  distilled duplicates,  distilled standards,
                  distilled spikes, and distilled blanks.

      8.4.4    When a steady reagent baseline is obtained and before
               starting the  sampler, adjust  the baseline  using the
               appropriate knob on  the colorimeter.  Aspirate a  calibration
               standard and  adjust  the STD CAL dial on the colorimeter until
               the desired signal  is obtained.  Record the STD CAL value.
               Reestablish the baseline and  proceed to analyze calibration
               standards, blanks, control standards, distilled samples,  and
               distilled QC  audits.

9.     Calculations

9.1   A separate determination of percent solids must be performed (see
      Part F).

9.2   The concentration of cyanide in the sample is determined as follows.

          9.2.1   (Titration)

                               (A - B)  x 	25° mL	 x 1000 g/kg
                                         mL aliquot titrated
                  CN,  mgAg	
                                               %solids
                                             X      '-- '
                                                 100
                                   D-85                    .            ILM01.0

-------
           WHERE:  A - mL of AgNOj for titration of sample
                         (1 mL - 1 mg Ag)
                   B - mL of AgNO-j for titraCion of blank
                         (1 mL - 1 mg Ag)
                   C — weC weight of original sample in g
                         (See 8.1.1)

           AND:  250 mL - volume of distillate (See 8.1.6)
                 1000 g/kg - conversion factor g to kg
                 mL aliquot titrated (See 8.2.1)
             % solids (see Part F)

9.2.2    (Manual  Spectrophotometric)
                                           100
            WHERE:  A  -  ug CN read from standard curve (per 250 mL)
                    B  -  mL of distillate taken for colorime'tric
                          determination (8.3.1)
                    C  -  wet weight of original sample in g
                          (See 8.1.1)

            The minimum value that can be substituted for A is 5
            ug/250 mL.  That yields a concentration of 10 ug/L in the
            distilled sample.

            AND:    50 mL - volume of standard taken for colorimetric
                            determination (See 8.3.1)
                    % solids (see Part F)

    9.2.3   (Semi -Automated Spectrophotometric)

            If the semi -automated method is used, measure the peak
            heights of the calibration standards (visually or using a
            data system) and calculate a linear  regression equation.
            Apply the equation to the samples and QC audits to
            determine the cyanide concentration  in the distillates.

                                    A x .25
                      CN,  mg/kg  "  c K % solids
                                           100
                             D-86                                ILM01.0

-------
                  WHERE:  A  -  ug/L determined from standard curve
                          C  -  wet weight of original sample in g
                                (See 8.1.1)

                  AND:    .25  -  conversion factor for distillate final
                                  volume (See 8.1.6)
                          % solids (see Part F)

                  The minimum value that can be substituted for A is 5
                  ug/250 mL.

Bibliography

1.    Modification of Method 335.2:   Cyanide, Total

2.    Methods for "Chemical Analysis of Water and Wastes",  March 1979.   EPA
      Publication #600/4-79-02.

3.    "Operation Manual for Technicon Auto Analyzer IIC System", 1980.
      Technical publication #TA9-0460-00.   Technicon Industrial Systems,
      Tarrytown, NY,  10591.

4.    "Users Guide for the Continuous Flow Analyzer Automation System",
      ESL, U.S. EPA,  Cincinnati,  Ohio (1981).

5.    "Interim Methods for the Sampling and Analysis of Priority Pollutants
      in Sediments and Fish Tissue,"  USEPA Environmental Monitoring and
      Support Laboratory,  Cincinnati,  Ohio,  August 1977,  Revised October
      1980.

6.    Op. cit.  (#5),  Methods  335.2,  modified (by  committee).
                                  D-87                                ILM01.0

-------
 ALLIHN CONDENSER	


 AIR INLET TUBE\
ONE LITER	
BOILING FLASK
— CONNECTING TUBING
                        GAS ABSORBER
                                              SUCTION
   Figure 1.  Cyanide  distillation apparatus
                    D-88
                      ILM01.0

-------
COOLING WATER
INLET
      HEATER -
                                       TO  LOW VACUUM
                                           SOURCE
                                   - ABSORBER
                              OISTILUHG FLASK
                   O
  Figure  2.   Cyanide distillation apparatus
                    D-89
ILM01.0

-------
sc  —i

o  <
                                           o
                                           o
                                        O  —
                                              fC
                                              o o
                         O  "C
                             C
                                              •e  o
                                              .o
                               «: ^-^
                               o
                               te  «->
                               o
              Figure  3.
Cyanide  Manifold

 D-90
                                                                  ILM01.0

-------
             METHOD FOR TOTAL CYANIDE  ANALYSIS  BY MIDI  DISTILLATION

                        CYANIDE,  TOTAL (water and soils)

             Method 335.2  CLP-M  (Semi-automated Spectrophotometric)

 1.     Scope and Application

 1.1   Cyanide determined by this method is  defined as  cyanide ion and complex
       cyanides converted to hydrocyanic acid by reaction in a reflux system
       with mineral acid in the  presence of  magnesium ion.

 1.2   This method covers  the determination  of  cyanide  by midi distillation
       with a semi-automated colorimetric analysis of the distillate.

 1.3   The detection limit for the semi-automated colorimetric method is
       approximately 10 ug/L.

 2.     Summary of Method

 2.1   The cyanide as  hydrocyanic  acid (HCN) is  released  from cyanide
       complexes by means  of a midi reflux-distillation operation  and absorbed
       in a scrubber containing  sodium hydroxide  solution.   The cyanide ion in
       the absorbing solution is  then  determined  colorimetrically.

 2.2   In the colorimetric measurement,  the  cyanide is  converted to  cyanogen
       chloride,  CNC1,  by  reaction with  chloramine-T at pH  less than 8 without
       hydrolysis to the cyanate.  After the reaction is  complete,  color  is
       formed on the addition of pyridinebarbituric acid  reagent.  The
       absorbance is read  at 580 nm.   To obtain colors  of comparable
       intensity,  it is essential  to have  the same  salt content in both the
       samples and the  standards.

 3.     Sample Handling  and Preservation

 3.1    All  bottles  must be  thoughly cleansed and rinsed to remove  soluble
       materials  from containers.

 3.2    Oxidizing agents  such as chlorine decompose most cyanides.   Test a drop
       of  the  sample with potassium iodide-starch test paper  (Kl-Starch
       paper); a blue color indicates the need for  treatment.  Add ascorbic
       acid, a few crystals at a time,  until a drop of sample produces no
       color on the indicator paper.  Then add additional 0.6 g of ascorbic
       acid for each liter of sample volume.

3.3    Samples are preserved with 2 mL of 10 N sodium hydroxide per liter of
       sample  (pH > 12) at the time of  collection.

3.4   Samples must be stored at  4°C (+2°C) and  must be  analyzed within the
      holding time specified in  Exhibit D, Section II.
                                   D-91                               ILM01.0

-------
4 .    Interferences

4.1   Interferences are eliminated or reduced by using the distillation
      procedure .

4.2   Sulfides adversely affect the colorimetric procedures.  If a drop of
      distillate on lead acetate test paper indicates the presence of
      sulfides, treat the sample with powdered cadmium carbonate.  Yellow
      cadmium sulfide precipitates if the sample contains sulfide.  Repeat
      this operation until a drop of the treated sample solution does not
      darken the lead acetate test paper.  Filter the solution through a dry
      filter paper into a dry beaker, and from the filtrate, measure the
      sample to be used for analysis .  Avoid a large excess of cadmium
      carbonate and long contact time in order to minimize loss by
      complexation or occlusion of cyanide on the precipitated material.

4.3   The presence of surfactants may cause the sample to foam during
      refluxing.   If this occurs, the addition of an agent such as Dow
      Corning 544 antifoaming agent will prevent the foam from collecting in
      the condenser.

5.    Apparatus

5.1   Midi reflux distillation apparatus as shown in figure 1.

5.2   Heating block - Capable of maintaining 125°C ±5°C.

5.3   Auto analyzer system with accessories:

      5.3.1    Sampler

      5.3.2    Pump

      5.3.3    Cyanide cartridge

      5.3.4    Colorimeter with 50 mm  flowcells and  580 nm  filter

      5.3.5    Chart recorder or  data  system.

5.4   Assorted volumetric glassware, pipets , and micropipets.

6.    Reagents

6.1   Distillation and Preparation Reagents

      6.1.1    Sodium hydroxide absorbing solution,  and sample wash solution,
               0.25  N.  Dissolve  10.0  g NaOH  in ASTM Type II water and dilute
               to  one liter.

      6.1.2    Magnesium  chloride solution,  51% (w/v) .  Dissolve  510 g of
                    6H20  in ASTM Type II water and  dilute  to one  liter.
      6.1.3    Sulfuric  acid,  50%  (v/v) .  Carefully add a portion  of
               concentrated t^SO^  to  an equal portion of ASTM  Type II water.

                                   D-92                                ILM01.0

-------
      6.1.4    Sodium hydroxide  solution, 1.25 N.  Dissolve  50  g of NaOH  in
               ASTM Type  II water and dilute to one liter.
6.2   Standards
      6.2.1    Stock cyanide solution, 1000 mg/L CN.  Dissolve  2.51  g of KCN
               and  2.0  g KOH in ASTM Type II water and dilute one  liter.
               Standardize with 0.0192 N AgN03 .

      6.2.2    Intermediate cyanide standard solution, 10 mg/L  CN.   Dilute  1.0
               mL of stock cyanide solution (6.2.1) plus 20 ml  of  1.25 N NaOH
               solution (6.1.4) to 100 mL with ASTM Type II water.   Prepare
               this solution at time of analysis.

      6.2.3    Rhodamine indicator.  Dissolve 20 mg of p-dimethylamino-benzal-
               rhodamine in 100 mL acetone.

      6.2.4    Silver nitrate solution, 0.0192 N.  Prepare by crushing
               approximately 5 g AgNO^ crystals and dry5~g to a constant
               weight at 104°C.  Weigh out 3.2647 g of dried AgN03 and
               dissolve in ASTM Type II water.  Dilute to one liter "( 1 mL
               corresponds to 1 mg CN) .
      6.2.5   Potassium chromate indicator solution.  Dissolve 50 g l^CRO^ In
              sufficient ASTM Type II water.  Add silver nitrate solution
              until a definite red precipitate is formed.  Let stand for at
              least 12 hours, filter, and dilute to one liter with ASTM Type
              II water.

      6.2.6   Primary standard sodium chloride, 0.0141 N.  Dissolve 824.1 mg
              NaCl (NBS- dried 20 minutes at 104°C) in ASTM Type II water and
              dilute to one liter.

      6.2.7   Sodium hydroxide solution, 0.1 N.  Dissolve 4 g of NaOH in ASTM
              Type II water and dilute to one liter.

6.3   Semi -Automated Spectrophotometric  Reagents

      6.3.1   Phosphate buffer solution, 1 M.  Dissolve 138 g of NaH2P04-H2O
              in ASTM Type II water and dilute to one liter.  Add 0.5 mL of
              Brij-35 (available from Technicon) . Store at 4°C .

      6.3.2   Chloramine-T solution, 0.4% (w/v) .  Dissolve 0.4 g of
              chloramine-T in ASTM Type II water and dilute to 100 mL.
              Prepare fresh at time of analysis.

      6.3.3   Color Reagent Solution, Pyridine barbituric acid color reagent
              solution.  Prepare this solution in the hood.  Transfer 15 g of
              barbituric acid into a one liter Erlenmeyer flask.  Add about
              100 mL of ASTM Type II water and swirl the flask Co mix.  Add
              75 mL of pyridine and 15 mL concentrated HCL and raix until all
              the barbituric acid is dissolved.  Dilute to one liter with
              ASTM Type II water and store at 4°C.


                                   D-93                                ILM01.0

-------
7.     Procedure

7.1   Distillation

      7.1.1   The procedure described here utilizes a midi distillation
              apparatus and requires a sample aliquot of 50 mLs or less for
              aqueous samples and one gram for solid materials.  NOTE:  All
              samples must initally be run undiluted (i.e., aqueous samples
              must first be run with a 50 mL aliquot and solid samples using
              a one gram sample) .  When the cyanide concentration exceeds the
              highest calibration standard, appropriate dilution (but not
              below the CRDL) and reanalysis of the sample is required.  The
              dilution factor must be reported on Form XIV.

      7.1.2   For aqueous samples:   Pipet 50 mL of sample,  or an aliquot
              diluted to 50 mL,  into the distillation flask along with 2 or 3
              boiling chips.

      7.1.3   For solid samples:   Weigh 1.0 g of sample (to the nearest 0.01
              g) into the distillation flask and dilute to  50 ml, with ASTM
              Type II water.   Add 2 or 3 boiling chips.

      7.1.4   Add 50 mL of 0.25  N NaOH (6.1.1)  to the  gas absorbing impinger.

      7.1.5   Connect the boiling flask,  condenser,  and absorber in the train
              as shown in figure  2.   The excess  cyanide trap  contains 0. 5  N
              NaOH.

      7.1.6   Turn on the vacuum and adjust the  gang (Whitney) values to give
              a flow of three bubbles per second from  the impingers in each
              reaction vessel.

      7.1.7   After five minutes  of vacuum flow,  inject 5 mL  of 50% (v/v)
              ^SO^ (6.1.3)  through the top air  inlet  tube  of the
              distillation head  into the reaction vessel.   Allow to mix for 5
              minutes.   (NOTE:   The acid volume  must be sufficient  to bring
              the sample/solution pH to below 2.0.)

      7.1.8   Add 2 mL of magnesium chloride solution  (6.1.2)  through the  top
              air inlet tube  of  the distillation head  into  the reaction
              flask.   Excessive  foaming from samples containing surfactants
              may be quelled  by  the addition of  another 2 mL  of magnesium
              chloride solution.

      7.1.9   Turn on the heating block and set  for  123-125°C.   Heat  the
              solution to boiling,  taking care  to prevent solution  backup  by
              periodic adjustment of the vacuum  flow.

      7.1.10   After one and a half  hours of refluxing,  turn off the heat and
              continue the vacuum for an additional  15  minutes.   The  flasks
              should be cool  at  this time.

      7.1.11   After cooling,  close  off the vacuum at the gang valve and
              remove  the absorber.   Seal the receiving  solutions  and  store

                                  D-94                                ILM01.0

-------
              them at 4°C until analyzed.  The solutions must be analyzed for
              cyanide within the 12 day holding time specified in Section II.

7.2   Semi-Automated Spectrophotometric Determination

      7.2.1   Operating conditions:  Because of the difference between
              various makes and models of satisfactory instruments, no
              detailed operating instructions can be provided.  The analyst
              should follow  the instructions provided by the manufacturer of
              the particular instrument.  It is the responsibility of the
              analyst to verify that the instrument configuration and
              operating conditions used satisfy the analytical requirements
              and to maintain quality control data confirming instrument
              performance and analytical results.

              The following general procedure applies to most semi-automated
              colorimeters.   Set up the manifold and complete system per
              manufacturer's instructions.   Allow the colorimeter and
              recorder warm up for at least  30 minutes prior to use.
              Establish a steady reagent baseline  feeding ASTM Type II water
              through the sample line and appropriate reagents (6.5)  through
              reagent lines.  Adjust the baseline  using the appropriate
              control on the colorimeter.

     7.2.2    Prepare a minimum of 3 standards and a blank by pipetting
              suitable volumes  of standard solution into 50 mL volumetric
              flasks.   NOTE:   One calibration standard must be at the
              Contract Required Detection Limit (CRDL).

              As  an  example,  standard solutions could be prepared as  follows:

                  Total ug CN
               standard solution         ml^lO mg/L CN       mL 0.05  N NaOH

                     0.00                   0.000                 20
                     0.10                   0.010                 20
                     0.25                   0.025                 20
                     0.50                   0.050                 20
                     1.00                   0.100                 20
                     2.00                   0.200                 20
                     5.00                   0.500                 20

              7.2.2.1    'Dilute  standards to  50 mL using ASTM Type II water.
                         It  is not  imperative  that all standards be distilled
                         in  the  same manner as  the samples.  At least one
                         standard  (mid-range) must be distilled and compared
                         to  similar values on  the curve  for each SDG to
                         ensure  the distillation technique  is reliable.   If
                         the distilled standard does not agree within +15% of
                         the undistilled standards, the operator must find
                         and correct the cause of the error before
                        proceeding.
                                  D-95                               ILM01.0

-------
      7.2.3    Aspirate  the  highest  calibration  standard and adjust  the
               colorimeter until  the desired  (maximum) signal-range  is
               obtained.

      7.2.4    Place calibration  standards, blanks, control standards in  the
               sampler tray,  followed by distilled samples, distilled
               duplicates, distilled standards,  distilled spikes,  and
               distilled blanks.

      7.2.5    Switch sample line from the ASTM  Type II water, to sampler, set
               the  appropriate  sampling rate and begin the analysis.

8.     Calculat ions

8.1   Calculations for Semi-automated Colorimetric Determination

      8.1.1    Prepare a standard curve by plotting absorbance (peak heights,
               determined visually or using a data system) of standards (y)
               versus cyanide concentration values (total ug CN/L) (x) .
               Perform a linear regression analysis.

      8.1.2    Multiply  all  distilled values by  the standardization value to
               correct for the  stock cyanide solution not being exactly 1000
               mg/L (See 6.2.1).

      8.1.3    Using the regression  analysis equation, calculate sample
               receiving solution concentrations from the calibration curve.

      8.1.4    Calculate the cyanide of aqueous  samples in ug/L of original
               sample, as follows:

                                          A x D x F
                            CM, ug/L  -       B


                 where:      A = ug/L CN of sample  from regression  analysis
                            B  —  Liter  of original  sample  for  distillation
                                 (0.050 L) (See  7.1.2)

                            D  =  any dilution  factor  necessary  to  bracket
                                  sample value within standard values

                            F - sample receiving solution volume  (0.050  L)

               The  minimum value  that can be substituted for A is  10 ug/L.

      8.1.5    Calculate the cyanide of solid samples in mg/kg of  original
               sample, as follows:

               8.1.5.1   A  separate determination of percent solids must be
                         performed  (See Part F) .
                                   D-96                                ILM01.0

-------
8.1.5.2    The concentration of cyanide in the sample is
           determined as follows:

                            A x D x F
             CN, ing/kg   -    B x E

           where:    A  -  ug/L  CN  of  sample  from  regression
                          analys is curve

                    B  - wet  weight  of  original  sample  in g
                          (See 7.1.3)

                    D  —  any   dilution  factor  necessary  to
                          bracket sample  value within standard
                          values

                    E - % solids (See Part F)/100.

                    F   -   sample  receiving   solution  volume
                          (0.050 L)

           The minimum value that  can be substituted  for  A is
                          10 ug/L
                    D-97                                ILM01.0

-------
               PART F -  PERCENT SOLIDS DETERMINATION PROCEDURE

 1.   Immediately following the weighing of the sample to be processed for
     analysis (see Section III, Part B- Soil/Sediment Sample Preparation),
     add 5-10 g of sample tro a tared weighing dish.  Weigh and record the
     weight to the nearest 0.01 g.

 2.   Place weighing dish plus sample, with the cover tipped to allow for
     moisture escape, in a drying oven maintained at 103-105°C.  Sample
     handling and drying should be conducted in a well-ventilated area.

 3.   Dry the sample overnight (12-24 hours) but no longer than 24 hours.  If
     dried less than 12 hours, it must be documented that constant weight
     was attained.*  Remove the sample from the oven and cool in a
     dessicator with the weighing dish cover in place before weighing.
     Weigh and record weight to nearest 0.01 g.  Do not analyze the dried
     sample.

 4.   Duplicate percent solids determinations are required at the same
     frequency as are other analytical determinations.   Duplicate results
     are to be recorded on FORM VI-IN.

 5.   For the duplicate percent solids determination, designate one sample
     aliquot as the "original" sample and the other aliquot as the
     "duplicate" sample.  Calculate dry weight using the results of the
     "original" sample aliquot.

 6.   Calculate percent solids by the formula below.   The value thus obtained
     will be reported on the appropriate FORM I-IN and,  where applicable,
     FORM VI-IN .   This  value will be used for calculating analytical
     concentration on a  dry weight basis.

                   % Solids  -   Sample Dry Weight x 10Q
                                    Sample Wet Weight
*For the purpose of paragraph 3,  drying time is  defined as the elapsed time
 in the oven; thus raw data must record time in and out of the oven to
 document the 12 hour drying time minimum.  In the event it is necessary to
 demons crate the attainmenr of constant: weight,  da~a masc be recorded for a
 minimum of two repetitive weigh/dry/dessicate/weigh cycles with a minimum
 of 1 hour drying time in each cycle.  Constant weight would be defined as
 a loss in weight of no greater than 0.01 g between the start weight and
 final weight of the last cycle.

                                   D-98                               ILM01.0

-------
           PART G - ALTERNATE METHODS ("CATASTROPHIC ICP FAILURE)4"
      Analyse                                                 Page No-.

      Aluminum - Method 202.2 CLP-M*, Furnace AA              D-101
      Barium - Method 208.2 CLP-M, Furnace AA                 D-102
      Cobalt - Method 219.2 CLP-M, Furnace AA                 D-103
      Copper - Method 220.2 CLP-M, Furnace AA                 D-104
      Iron - Method 236.2 CLP-M, Furnace AA                   D-105
      ^anganese - Method 243.2 CLP-M, Furnace AA              D-106
      Nickel - Method 249.2 CLP-M, Furnace AA                 D-107 -
      Vanadium - Method 286.2 CLP-M, Furnace AA               D-108
      Zinc - Method 289.2 CLP-M, Furnace AA                   D-109
      Aluminum - Method 202.1 CLP-M, Flame AA                 D-lll
      Antimony - Method 204.1 CLP-M, Flame AA                 D-113
      Barium - Method 208.1 CLP-M, Flame AA                   D-114
      Beryllium - Method 210.1 CLP-M, Flame AA                D-115
      Cadmium - Method 213.1 CLP-M, Flame AA                  D-116
      Chromium - Method 218.1 CLP-M, Flame AA                 D-117
      Cobalt - Method 219.1 CLP-M, Flame AA                   D-118
      Copper - Method 220.1 CLP-M, Flame AA                   D-119
      Iron - Method 236.1 CLP-M, Flame AA                     D-120
      Lead - Method 239.1 CLP-M, Flame AA                     D-121
      Manganese - Method 243.1 CLP-M, Flame AA                D-122
      Nickel - Method 249.1 CLP-M, Flame AA                   D-123
      Silver - Method 272.1 CLP-M, Flame AA                   D-125
      Thallium - Method 279.1 CLP-M, Flame AA                 D-126
      Vanadium - Method 286.1 CLP-M, Flame AA                 D-127
      Zinc - Method 289.1 CLP-M, Flame AA                     D-128
 Furnace AA Methods are from "Methods for Chemical Analysis of Water and
 Wastes", (EPA-600/4-79-02), March 1979, as modified for use  in  the
 Contract Laboratory Program (CLP).  Flame AA (Flame Technique)  Methods are
 from "Interim Methods for the Sampling and Analysis of Priority Pollutants
 in Sediments and Fish Tissue," USEPA Environmental Monitoring and Support
 Laboratory, Cincinnati, Ohio, August 1977, Revised October 1980, as
 modified for use in the CLP.

*CLP-M  Modified for the Contract Laboratory Program.
                                   D-99                                ILM01.0

-------
                   CONDITIONS  FOR USE OF ALTERNATE METHODS


The methods contained in Part G may be used only if all of the following
conditions are met:

     1)    Catastrophic failure of ICP occurs,

     2)    Administrative Project Officer authorization  for  use  of alternate
          methods  is  granted,  and

     3)    The  IDLs for the  instrumentation have been determined,  as per
          Exhibit  E,  within the current calendar quarter.
                                  D-100                                ILM01.0

-------
                                  ALUMINUM*

         Method 202.2 CLP-M** (Atomic Absorption, Furnace Technique)


Optimum Concentration Range:   20-200 ug/L
Approximate  Detection Limit:   3  ug/L

Preparation  of  Standard Solution

1.   Stock solution:  Prepare as described under AA Flame Technique  (Method
     202.1 CLP-M).

2.   Prepare dilutions  of the stock  solution to be used as calibration
     standards at the time of analysis.  These solutions are also to be used
     for "standard additions".

3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying Time and Temp:   30 sec <§ 125°C.
      2.  Ashing Time and Temp:   30 sec @ 1300°C.
      3.  Atomizing Time and Temp:   10 sec  @ 2700°C.
      4.  Purge  Gas Atmosphere:   Argon
      5.  Wavelength:   309.3 nm
      6.  Other  operating parameters should be  set as  specified by the
          particular instrument manufacturer.
Notes

1.   The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100,  based on the use of a 20 uL injection,
     continuous  flow purge gas and non-pyrolytic graphite and are to be used
     as guidelines only.

2.   Background  correction is required.

3.   It has been reported that chloride ion and that nitrogen used as a
     purge gas suppress the aluminum signal.   Therefore the use of halide
     acids and nitrogen as a purge gas should be avoided.

4.   For every sample analyzed,  verification is necessary to determine that
     method of standard addition is not required (Exhibit E).

5.   If method of standard addition is required,  follow the procedure given
     in Exhibit  E.
-'This method may only be used under specified conditions.

**CLP-M  Modified for the Contract Laboratory Program.

                                   D-101                                ILM01.0

-------
                                   BARIUM*

         Method  208.2  CLP-M**  (Atomic Absorption,  Furnace Technique)

Optimum Concentration Range:  10-200 ug/L
Approximate Detection Limit:  2 ug/L

Preparation of Standard Solution

1.   Stock solution:  Prepare  as described  under AA Flame Technique (Method
     208.1 CLP-M).

2.   Prepare dilutions of the  stock solution to be used as calibration
     standards at the time of  analysis.   These solutions are also to be used
     for "standard additions".

3.   The calibration standards  must be prepared using the same type of acid
     and at the  same concentration as will  result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying Time  and Temp:   30  sec @ 125°.
      2.  Ashing Time  and Temp:   30  sec @ 1200°C.
      3.  Atomizing Time  and Temp:   10 sec  @ 2800°C.
      A.  Purge  Gas Atmosphere:   Argon
      5.  Wavelength:   553.6 run
      6.  Other  operating parameters should be set as specified by the
          particular instrument manufacturer.
Notes

1.   The above concentration values and  instrument conditions are for a
     Perkin-Elmer HGA-2100,  based on the  use of a 20 uL injection,
     continuous  flow purge gas  and pyrolytic graphite and are to be used as
     guidelines  only.

2.   The use of  halide acid should be avoided.

3.   Because of  possible  chemical interaction, nitrogen should not be used
     as a purge  gas.

4.   For every sample analyzed, verification is necessary to determine that
     method of standard addition is not  required (see Exhibit E).

5.   If method of standard addition is required,  follow the procedure given
     in Exhibit  E.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.
                                  D-102                                ILM01.0

-------
                                   COBALT*

         Method 219.2  CLP-M**  (Atomic Absorption,  Furnace Technique)

Optimum Concentration Range:  5-LOO ug/L
Approximate Detection Limit:  1 ug/L

Preparation of Standard Solution

1.   Stock solution:  Prepare as described under AA Flame Technique (Method
     219.1 CLP-M).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  These solutions are also to be used
     for "standard additions".

3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying Time  and Temp:   30 sec @ 125°C.
      2.  Ashing-Time  and Temp:   30 sec @ 900°C.
      3.  Atomizing Time  and Temp:   10 sec @ 2700°C.
      4.  Purge Gas Atmosphere:   Argon
      5.  Wavelength:   240.7 run
      6.  Other operating parameters should be set as specified by the
          particular instrument manufacturer.

Notes

1.   The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100,  based on the use of a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite and are to be u--d
     as guidelines  only.   Smaller size furnace devices or those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter  time periods than the above recommended
     settings.

2.   The use of background correction is required.

3.   Nitrogen may also be used as the purge gas but with reported low
     sensitivity.

4.   For every sample analyzed,  verification is necessary to determine that
     method of standard addition is not required (see Exhibit E).

5.   If method of standard addition is required,  follow the procedure given
     in Exhibit E.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                   D-103                                ILM01.0

-------
                                   COPPER*

         Method 220.2 CLP-M** (Atomic Absorption,  Furnace Technique)

Optimum Concentration Range:  5-100 ug/L
Approximate Detection Limit:  1 ug/L

Preparation of Standard  Solution

1.   Stock solution:  Prepare as described under AA Flame Technique (Method
     220.1 CLP-M).

2.   Prepare dilutions of che stock solution to be used as calibration
     standards at the time of analysis.  These solutions are also to be used
     for "standard additions".

3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying Time and Temp:   30 sec @ 125°C.
      2.  Ashing Time and Temp:   30 sec @ 900°C.
      3.  Atomizing Time and Temp:   10 sec @ 2700°C.
      A.  Purge Gas Atmosphere:   Argon
      5.  Wavelength:   324.7 nm
      6.  Other operating parameters  should be  set as  specified by the
          particular instrument  manufacturer.
Notes

1.   The above concentration values and instrument conditions  are for a
     Perkin-Elmer HGA-2100,  based on the use of a  20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite and  are to  be used
     as guidelines only.  Smaller size furnace  devices or those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter time periods than the above recommended
     settings.

2.   Background correction is required.

3.   Nitrogen  may also be used as the purge gas.

4.   For every sample analyzed,  verification is necessary to determine  that
     method of standard addition is not required (see  Exhibit  E).

5.   If method of standard addition is required, follow the  procedure  given
     in Exhibit E.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-104                               ILM01.0

-------
                                     IRON*

         Method 236.2  CLP-M**  (Atomic Absorption, Furnace Technique)

 Optimum Concentration Range:  5-100 ug/L
 Approximate Detection Limit:  1 ug/L

 Preparation of Standard Solution

 1.    Stock solution:   Prepare  as described under AA Flame Technique  (Method
      236.1 CLP-M).

 2.    Prepare dilutions of the  stock solution to  be  used  as  calibration
      standards at  the  time of  analysis.  These solutions are also  to be used
      for "standard additions".

 3.    The calibration  standards  must be prepared  using  the same  type of acid
      and at the same  concentration as will result in the sample  to be
      analyzed after sample preparation.

 Instrument Parameters (General)

       1.  Drying Time  and Temp:   30 sec @  125°C.
       2.  Ashing Time  and Temp:   30 sec @  1000°C.
       3.  Atomizing Time  and Temp:   10 sec @ 2700°C.
       4.  Purge Gas Atmosphere:   Argon
       5.  Wavelength:   248.3 nm
       6.  Other operating parameters should be set as specified by the
           particular instrument manufacturer.


 1.    The above concentration values and instrument conditions are for a
      Perkin-Elmer  HGA-2100, based on the use of  a 20 uL  injection,
      continuous flow purge gas  and non-pyrolytic graphite and are to be used
      as  guidelines  only.   Smaller size furnace devices or those employing
      faster  rates  of atomization  can be operated using lower atomization
      temperatures  for  shorter time  periods than  the above recommended
      settings.

2.   The use of background correction is required.

3.   Nitrogen may also be used as  the purge gas.

4.   For every sample analyzed, verification is necessary to determine  that
     method of standard addition  is not required  (see Exhibit E).

5.   If method of standard addition is required,  follow the  procedure given
     in Exhibit E.
*This method may only be used under specified conditions.

        Modified for the Contract Laboratory Program.

                                   D-105                               ILM01.0

-------
                                 MANGANESE*
        Method  243.2 CLP-M**  (Atomic Absorption, Furnace  Technique)

Optimum Concentration Range:   1-30 ug/L
Approximate Detection Limit:   0.2 ug/L

Preparation of Standard Solution

1.   Stock solution:   Prepare  as described  under AA Flame  Technique (Method
     243.1 CLP-M).

2.   Prepare dilutions  of the  stock solution  to  be  used as calibration
     standards at the time of  analysis.   These solutions are also to be used
     for "standard additions".

3.   The calibration standards must be  prepared  using the  same type of acid
     and at the  same concentration as will  result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying Time and Temp:   30 sec @ 125°C.
      2.  Ashing Time and Temp:   30 sec @ 1000°C.
      3.  Atomizing Time  and Temp:   10  sec  @  2700°C.
      4.  Purge  Gas Atmosphere:   Argon
      5.  Wavelength:   279.5 nm
      6.  Other  operating parameters should be set  as specified by the
          particular instrument manufacturer.

Notes
1.   The above concentration values and instrument  conditions are for a
     Perkin-Elmer HGA-2100, based on the  use  of  a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite  and are to be used
     as guidelines only.   Smaller size  furnace devices or  those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter  time periods  than  the above  recommended
     settings.

2.   The use of background correction is  required.

3.   Nitrogen may also be used as the  purge gas.

4.   For every sample analyzed, verification  is  necessary  to determine that
     method of standard addition is not required (see Exhibit E) .

5.   If method of standard addition is  required,  follow the procedure given
     in Exhibit E.
*This method may only be used under specified conditions.

•'""CLP-M Modified for the Contract Laboratory Program.
                                   D-106                                ILM01.0

-------
                                   NICKEL*

         Method 249.2  CLP-M** (Atomic  Absorption,  Furnace  Technique)

Optimum Concentration Range:  5-100 ug/L
Approximate Detection Limi c:  1 ug/L

Preparation of Standard Solution

1.   Stock solution:   Prepare as described under AA Flame Technique (Method
     249.1 CLP-M).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the  time of analysis.  These solutions are also to  be used
     for "standard additions".

3.   The calibration standards  must be prepared using the same type of acid
     and at the same concentration as  will result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying Time  and Temp:   30 sec @  125°C.
      2.  Ashing Time  and Temp:   30 sec 
-------
                                  VANADIUM*
         Method 286.2 CLP-M** (Atomic Absorption, Furnace Technique)

Optimum  Concentration Range:   10-200 ug/L
Approximate Detection Limit:   A ug/L

Preparation of Standard Solution

1.   Stock solution:  Prepare as described under AA Flame Technique (Method
     286.1 CLP-M).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  These solutions are also to be used
     for "standard additions."

3.   The calibration standards must be prepared using the same type of acid
     and at the same concentration as will result in the sample to be
     analyzed after sample preparation.

Instrument Parameters (General)

      1.  Drying  Time and  Temp:  30 sec @ 125°C.
      2.  Ashing  Time and  Temp:  30 sec @ 1400°C.
      3.  Atomizing Time and Temp:   15  sec  @ 2800°C.
      A.  Purge Gas Atmosphere:  Argon
      5.  Wavelength:  318.4 run
      6.  Other operating  parameters should be  set as  specified by the
          particular instrument manufacturer.

Notes

1.   The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100.  based on the use of a 20 uL injection,
     continuous flow purge gas and pyrolytic graphite and are to be used as
     guidelines only.  Smaller size furnace devices or those employing
     faster rates of atomization can be operated vising lower atomization
     temperatures for shorter time periods than the above recommended
     settings.

2.   The use  of background correction is required.

3.   Because  of possible  chemical interaction,  nitrogen should not be used
     as  the purge gas.

4.   For every sample analyzed, verification is necessary to determine  that
     method of standard addition is not required (see Exhibit E).

5.   If  method of standard addition is  required,  follow the  procedure given
     in  Exhibit E.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-108                               ILM01.0

-------
                                    ZINC*

 Method 289.2  CLP-M** (Atomic Absorption,  Furnace Technique)

 Optimum Concentration Range:  0.2-4 ug/L
 Approximate Detection Limit:  0.05 ug/L

 Preparation of Standard Solution
 1.   Stock solution:   Prepare  as  described  under AA Flame  Technique  (Method
     289.1 CLP-M).

 2.   Prepare dilutions  of  the  stock solution  to  be  used as calibration
     standards at the  time  of  analysis.   These solutions are  also to be used
     for "standard additions".

 3.   The calibration standards must be prepared  using  the  same type of acid
     and at the same concentration as will  result in the sample to be
     analyzed after sample  preparation.

 Instrument Parameters  (General^
       1.  Drying Time and Temp:  30 sec @ 125°C.
       2.  Ashing Time and Temp:  30 sec @ 400°C.
       3.  Atomizing Time and Temp:  10 sec @ 2500°C.
       4.  Purge Gas Atmosphere:  Argon
       5.  Wavelength:  213.9 nm
       6.  Other operating parameters should be set as specified by the
          particular instrument manufacturer.
     The above concentration values and instrument conditions are for a
     Perkin-Elmer HGA-2100, based on the use of a 20 uL injection,
     continuous flow purge gas and non-pyrolytic graphite and are to be used
     as guidelines only.  Smaller size furnace devices or those employing
     faster rates of atomization can be operated using lower atomization
     temperatures for shorter time periods then the above recommended
     settings.

2.   The use of background correction is required.

3.   Nitrogen may also be used as the purge gas.

4.   The analysis of zinc by the graphite furnace is extremely sensitive and
     very subject to 'contamination from the work area,  reagents,  and pipette
     tips.   Since all these factors affect the precision and accuracy,  zinc
     should be  analyzed by the direct aspiration procedure whenever
     possible.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-109                               ILM01.0

-------
5.    For every sample analyzed,  verification is necessary  to  determine that
     method of standard addition is  not  required  (see  Exhibit E).

6.    If method of standard addition  is required,  follow  the procedure  given
     in Exhibit E.
                                  D-110                                ILM01.0

-------
                                  ALUMINUM*

          Method 202.1  CLP-M** (Atomic Absorption,  Flame Technique)


Optimum Concentration Range:  5-50 mg/L using a wavelength of 309.3 nm
Sensitivity:  1 mg/L
Approximate Detection Limit:  0.1 mg/L

Preparation of Standard Solution

1.   Stock Solution:  Carefully weigh 1,000  g of aluminum  metal  analytical
     reagent grade).  Add 15 mL of cone.   HC1 and 5 mL cone.   HNO^ to the
     metal, cover the beakec and warm gently.  When solution  is  complete,
     transfer quantitatively to a liter volumetric flask and  make up to
     volume with deionized distilled water.      1 mL - 1 mg Al (1000 mg/L).

2.   Potassium Chloride Solution:  Dissolve  95 g potassium chloride (KC1)  in
     deionized distilled water and make up to 1 liter.

3.   Prepare dilutions of che stock solution to be used as calibration
     standards at the time of analysis.   The calibration standards must be
     prepared using the same type of acid and at the same  concentration as
     will result in the sample to be analyzed after sample preparation.  To
     each 100 mL of standard and sample alike add 2.0 mL potassium chloride
     solution.

Instrument Parameters  (General)

      1.  Aluminum hollow cathode lamp
      2.  Wavelength:   309.3 nm
      3.  Fuel:   Acetylene
      4.  Oxidant:   Nitrous oxide
      5.  Type of flame:  Fuel rich
Interferences

1.   Aluminum is partially ionized in the nitrous oxide-acetylene flame.
     This problem may be controlled by the addition of an  alkali metal
     (potassium, 1000 ug/mL) to both sample  and standard solutions.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-lll                                ILM01.0

-------
Notes

 1.  The following may also be used:

       308.2 nm Relative Sensitivity 1
       396.2 nm Relative Sensitivity 2
       394.4 nm Relative Sensitivity 2.5

 2.  For concentrations of aluminum below 0.3 mg/L, use of
     Furnace Technique (Method 202.2 CLP-M) is recommended.
                                  D'112                                ILM01.0

-------
                                  ANTIMONY*

          Method 204.1 CLP-H** (Atomic Absorption, Flame Technique)


 Optimum  Concentration Range:   1-40  mg/L using  a  wavelength of 217.6  nm
 Sensitivity:   0.5  mg/L
 Approximate Detection Limit::   0.2 mg/L

 Preparation of Standard  Solution

 1.   Stock Solution:   Carefully weigh  2.7426 g of antimony potassium
     tartrate  (analytical  reagent grade) and dissolve in deionized distilled
     water.  Dilute to 1 liter with deionized distilled water.  1 mL - 1 mg
     Sb  (1000 mg/L).

 2.   Prepare dilutions of  rhe stock solution to be used as calibration
     standards at the  time of analysis.  The calibration standards must be
     prepared using the same type of acid and at the same concentration as
     will result in the sample to be analyzed after sample preparation.

 Instrumental Parameters  (General)

      1.   Antimony hollow cathode lamp
      2.   Wavelength:   217.6  nm
      3.   Fuel:  Acetylene
      4.   Oxidant:   Air
      5.   Type  of flame:   Fuel lean

 Interferences

 1.   In the presence of lead (1000 mg/L), a special interference may occur
     at the 217.6 nm resonance line.  In this case the 231.1 nm antimony
     line should be used.

 2.   Increasing acid concentrations decrease antimony absorption.   To avoid
     this" effect, the  acid concentration in the samples  and in the standards
     must be matched.

 Notes

 1.   For  concentrations of antimony below 0.35  mg/L,  use of the Furnace
     Technique  (Method 204.2  CLP-M)  is recommended.
•"This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-113                               ILM01.0

-------
                                   BARIUM*

         Method 208.1  CLP-M** (Atomic Absorption, Flame Technique)

Optimum Concentration Range:  1-20 mg/L using a wavelength of 553.6 nm
Sensitivity:  0.4 mg/L
Approximate Detection Limic:  0.1 mg/L

Preparation of Standard Solution

1.   Stock Solution:  Dissolve 1.7787 g of barium chloride (BaCl2'2H20,
     analytical reagent grade) in deionized distilled water and dilute to
     liter.  1 mL - 1 mg Ba (1000 mg/L).

2.   Potassium chloride solution:  Dissolve 95 g potassium chloride,  KC1, in
     deionized distilled water and make up to 1 liter.

3.   Prepare dilutions of r.he stock solution to be used as calibration
     standards at the time of analysis.   To each 100 mL of standard and
     sample alike add 2.0 mL potassium chloride solution.   The calibration
     standards must be prepared using the same type  of ac^d and at the same
     concentration as will result in the sample to be analyzed after sample
     preparation.

Instrumental Parameters (General)

      1.  Barium hollow cathode  lamp
      2.  Wavelength:   553.6 nm
      3.  Fuel:   Acetylene
      4.  Oxidant:   Nitrous  oxide
      5.  Type of flame:   Fuel rich
Interferences

1.   The use of a nitrous oxide-acetylene flame virtually  eliminates
     chemical interference;  however, barium is easily ionized in this flame
     and potassium must be added (1000 mg/L) to standards  and samples alike
     to control this effect.

2.   If the nitrous oxide flame is not available and acetylene-air is used,
     phosphate,  silicon and aluminum will severely depress the barium
     absorbance.  This may be overcome by the addition of  2000 mg/L
     lanthanum.

Notes

1.   For concentrations of barium below 0.2 mg/L, use of the Furnace
     Technique (Method 208.2 CLP-M) is recommended.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                   D-114                               ILM01.0

-------
                                 BERYLLIUM*

          Method 210.1 CLP-M** (Atomic Absorption,  Flame  Technique)


Optimum Concentration Range:  0.052 mg/L using a wavelength of 234.9 nm
Sensitivity:  0.025 mg/L
Approximate Detection Limit:  0.005 mg/L

Preparation of Standard Solution

1.   Stock Solution:  Dissolve 11.6586 g of beryllium sulfate, BeSO^,  in
     deionized distilled water containing 2 mL cone, nitric acid and dilute
     to 1 liter.  1 mL - 1 mg Be (1000 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  The calibration standards should be
     prepared using the same type of acid and at the same concentration as
     will result in the sample to be analyzed after sample preparation.

Instrumental Parameters (General)

      1.  Beryllium hollow cathode lamp
      2.  Wavelength:   234.9 nm
      3.  Fuel:   Acetylene
      4.  Oxidant:   Nitrous oxide
      5.  Type of flame:   Fuel rich

Interferences

1.   Sodium and silicon at concentrations in excess of 1000 mg/L have  been
     found to severely depress the beryllium absorbance.

2.   Bicarbonate ion is reported to interfere;  however,  its effect is
     eliminated when samples are acidified to a pH of 1.5.

3.   Aluminum at concentrations of 500 ug/L i~  reported to depress.the
     sensitivity of beryllium [Spectrochim Acta 22, 1325  (1966)].

Notes

1.   For concentrations of beryllium below 0.02 mg/L,  use of the  Furnace
     Technique (Method 210.2 CLP-M) is recommended.
*This method may only be used under specified conditions.

-—-CLP-M. Modified for the Contract Laboratory Program.

                                  D-115                               ILM01.0

-------
                                   CADMIUM*

          Method 213.1 CLP-M** (Atomic Absorption, Flame Technique)


 Optimum Concentration Range:   0.052  mg/L using a  wavelength of  228.8  ran
 Sensitivity:   0.025  mg/L
 Approximate  Detection Limic:   0.005  mg/L

 Preparation  of Standard  Solution

 1.   Stock Solution:   Carefully weigh 2.282 g  of  cadmium sulfate
     (3CdSO^'8H20, analytical reagent grade) and  dissolve in deionized
     distilled water.  Make up to 1  liter with dionized distilled water.
     1 mL - 1  mg Cd  (1000 mg/L).

 2.   Prepare dilutions of the stock  solution to be used as calibration
     standards at the  time of analysis.  The calibration standards must be
     prepared  using  the  same type of acid and at  the same concentration as
     will result in  the  sample to be analyzed after sample preparation.

 Instrumental Parameters  (General)

      1.  Cadmium hollow cathode lamp
      2.  Wavelength:  228.8 nm
      3.  Fuel:  Acetylene
      4.  Oxidant:  Air
      5.  Type of flame:   Oxidizing

 Notes

 1.   For concentrations  of cadmium below 20 ug/L,  use of the Furnace
     Technique, Method 213.2 CLP-M is recommended.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-116                               ILM01.0

-------
                                  CHROMIUM*

          Method 218.1 CLP-M** (Atomic Absorption, Flame Technique)

 Optimum Concentration Range:   0.5-10  mg/L using a wavelength of 357.9 nra
 Sensitivity:   0.25 mg/L
 Approximate  Detection Limi t:   0.05  tng/L

 Preparation  of Standard Solution

 1.   Stock Solution:   Dissolve 1.923  g  of chromium trioxide  (CrO^,  reagent
     grade) in deionized distilled water.  When solution is  complete,
     acidify with  redistilled HNO-j and  dilute to  1 liter with deionized
     distilled water.  1 mL - 1  mg Cr (1000 tng/L).
 2.   Prepare dilutions of the stock solution to be used as calibration
     standards  at  the  time  of analysis.   The calibration standards  must be
     prepared  using the same type of  acid and at  the same concentration as
     will result in the sample to be  analyzed after sample preparation.

 Instrument Parameters  (General)

          Chromium hollow cathode lamp
          Wavelength:  357.9 nm
          Fuel:  Acetylene
          Oxidant:   Nitrous oxide
          Type of flame:  Fuel rich


 1.   The following wavelengths may also be used:

          359.3 nm Relative Sensitivity 1.4
          425.4 nm Relative Sensitivity 2
          427.5 nm Relative Sensitivity 3
          428.9 nm Relative Sensitivity 4

 2.   The fuel  rich air-acetylene flame provides greater sensitivity but is
     subject to chemical and matrix interference from iron, nickel, and
     other metals.   If the  analysis is performed in a lean flame the
     interference can be lessened but the sensitivity will also be reduced.

 3.   The suppression of both Cr  (III)  and Cr (VI) absorption by most
     interfering ions in fuel rich air-acetylene flames is reportedly
     controlled by the addition of 1%  ammonium bifluoride in 0.2% sodium
     sulfate [Talanta 20, 631 (1973)].  A 1% oxine solution is also reported
     to be useful.

4.   For concentrations of chromium between 50 and 200 ug/L where the air-
     acetylene flame cannot be used or for concentrations below 50 ug/L,  use
     of the Furnace Technique (Method  218.2 CLP-M) is  recommended.
"'This method may only be used under specified conditions .

**CLP-M Modified for the Contract Laboratory Program.

                                  D-117                                ILM01.0

-------
                                   COBALT*

         Method  219.1** CLP-M (Atomic Absorption,  Flame Technique)


Optimum Concentration Range:  0.5-5 mg/L using a wavelength of 240.7 nm
Sensitivity:  0.2 mg/L
Approximate Detection Limit:  0.05 mg/L

Preparation of Standard Solution

1.   Stock Solution:   Dissolve 4.307 g of cobaltous chloride (CoC12.   6H20
     analytical reagent grade),  in deionized distilled water.   Add 10 mL of
     concentrated nitric acid and dilute to 1 liter with deionized distilled
     water.   1 mL - 1 mg Co (1000 mg/L).

2.   Prepare dilutions of the stock cobalt  solution to be  used as
     calibration standards at the time of analysis.  The calibration
     standards must be prepared using the same type of acid and at the same
     concentration as will result in the sample to be analyzed after sample
     preparation.

Instrument Parameters (General)

      1.  Cobalt hollow cathode  lamp
      2.  Wavelength:   240.7 nm
      3.  Fuel:  Acetylene
      4.  Oxidant:  Air
      5.  Type of  flame:   Oxidizing
1.   For concentrations of cobalt below 100 ug/L use of the Furnace
     Technique (Method 219.2 CLP-M)  is recommended.
*This method may only be vised under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                   D-118                                ILM01.0

-------
                                   COPPER*

         Method 220.1  CLP-H** (Atomic Absorption,  Flame Technique)


Optimum Concentration Rang^:   0.2-5 mg/L using a wavelength of 324.7 nrn
Sensitivity:  0.1 mg/L
Approximate Detection Limit:   0.02 rag/L

Preparation of Standard Solution

1.   Stock Solution:  Carefully weigh 100 g of electrolyte copper
     (analytical reagent grade).   Dissolve in 5 mL redistilled HNOo and make
     up to 1 liter with deionized distilled water.  Final concentration is 1
     mg Cu per mL (1000 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  The calibration standards must be
     prepared using the same  type of acid and at the same concentration as
     will result in the sample to be analyzed after sample preparation.

Instrumental Parameters (General)

          Copper hollow ca chode lamp
          Wavelength:   324.7  run
          Fuel:   Acetylene
          Oxidant:   Air
          Type of flame:   Oxidizing
1.   For concentrations of copper below 50 ug/L use of the Furnace Technique
     (Method 220.2 CLP-M) is recommended.

2.   Numerous absorption lines are available for the determination of
     copper.  By selecting a suitable absorption wavelength,  copper samples
     may be analyzed over a very wide range of concentrations.  The
     following lines may be used:
          327.4 nm Relative Sensitivity 2
          216.5 nm Relative Sensitivity 7
          222.5 nm Relative Sensitivity 20
*This method may only be used under specified conditions.

**cLP-M Modified for che Contract Laboratory Program.

                                   D-119                                ILM01.0

-------
                                    IRON*

          Method 236.1 CLP-M** (Atomic Absorption, Flame Technique)


 Optimum Concentration Range:   0.3-5 mg/L using  a wavelength of 248.3  ran
 Sensitivity:   0.12 mg/L
 Approximate  Detection Limit:   0.03 mg/L

 Preparation  of Standard Solution

 1.   Stock Solution:   Carefully  weigh 1.000 g of pure iron wire (analytical
     reagent  grade)  and  dissolve in 5 mL redistilled HNO-j, wanning if
     necessary.  When solution is complete, make up to 1 liter with
     deionized  distilled water.  1 mL -  1 mg Fe  (1000 mg/L).

 2.   Prepare  dilutions of  the stock solution to be used as calibration
     standards  at  the time of analysis.  The calibration standards must be
     prepared using  the  same type of acid and at the same concentration as
     will result in  the  sample to be analyzed after sample preparation.

 Instrumental  Parameters  (General)

       1.  Iron hollow cathode lamp
       2.  Wavelength:  248.3 nm
       3.  Fuel:  Acetylene
       4.  Oxidant:   Air
       5.  Type of flame:  Oxidizing

 Notes

 1.   The following wavelengths may also be used:
     248.8 nm Relative Sensitivity 2
     271.9 nm Relative Sensitivity 4
     302.1 nm Relative Sensitivity 5
     252.7 nm Relative Sensitivity 6
     372.0 nm Relative Sensitivity 10

 2.   For concentrations  of iron below 0.05 mg/L use of the Furnace Technique
     (Method  236.2 CLP-M) is recommended.
•*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-120                               ILM01.0

-------
                                    LEAD*

          Method 239.1 CLP-M** (Atomic Absorption, Flame Technique)


 Optimum Concentration Range:   1-20  mg/L using  a wavelength of 283.3 nra
 Sensitivity:   0.5  mg/L
 Approximate Detection Limit:   0.1 mg/L

 Preparation of Standard  Solution

 1.    Stock  Solution:   Carefully  weigh  1.599 g  of  lead nitrate,  Pb(N03)2
      (analytical reagent grade), and dissolve  deionized distilled  water.
      When solution is  complete acidify with 10 mL redistilled HNO^ and
      dilute to  1 liter with deionized  distilled water.  1  mL  - 1 mg Pb (1000
      mg/L).

 2.    Prepare dilutions of  the stock solution to be used as  calibration
      standards  at  the  time of analysis.  The calibration standards must be
      prepared using the same type of acid and  at  the same  concentration as
      will result in the sample to be analyzed  after sample  preparation.

 Instrumental Parameters  (General)

       1.  Lead hollow  cathode lamp
       2.  Wavelength:  283.3 run
       3,  Fuel:  Acetylene
       4.  Oxidant:   Air
       5.  Type of  flame:   Oxidizing
 1.   The  analysis of this metal is exceptionally sensitive to  turbulence and
     absorption bands in the flame.  Therefore, some care should be taken to
     position the light beam in the most stable, center portion of the
     flame.  To do this, first adjust the burner to maximize the absorbance
     reading with a lead standard.  Then, aspirate a water blank and make
     minute adjustments in the burner alignment to minimize the signal.

 2.   The  concentrations of lead below 200 ug/L use of the Furnace Technique
     (Method 239.2 CLP-M) is recommended.

 3.   The  following wavelengths may also be used:

         217.0 nm  Relative Sensitivity  0.4
         261.4 nm  Relative Sensitivity  10
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-121                                ILM01.0

-------
                                 MANGANESE*

         Method  243.1  CLP-M**  (Atomic Absorption,  Flame  Technique)


Optimum Concentration Ran^e:   0.1-3 mg/L using a wavelength of 279.5 nm
Sensitivity:  0.05 mg/L
Approximate Detection Limit:   0.01 mg/L

Preparation of Standard Solution

1.   Stock Solution:   Carefully weigh 1.000 g of manganese metal  (analytical
     reagent grade),  and dissolve in 10  mL redistilled HNO-j.   When solution
     is complete, dilute' to 1 liter with 1% (v/v) HC1.   1 mL  - 1  mg Mn (1000
     mg/L).
2.   Prepare dilutions of ch» stock solution to be used as calibration
     standards at the time of analysis.   The calibration standards must be
     prepared using the same type of acid and at the same concentration as
     will result in the sample to be analyzed after sample preparation.

Instrumental Parameters (General)

      1.  Manganese hollow cathode lamp
      2.  Wavelength:  279.5 nm
      3.  Fuel:   Acetylene
      4.  Oxidant:   Air
      5.  Type of flame:   Oxidizing
Notes

1.   For concentrations of manganese below 25 ug/L, use of the Furnace
     Technique (Method 243.2 CLP-M) is recommended.

2.   The following line may also be used:  403.1 nm Relative Sensitivity 10.
"'This method may only be ui,ed under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                   D-122                                ILM01.0

-------
                                   NICKEL*

          Method 249.1 CLP-M** (Atomic Absorption, Flame Technique)


Optimum  Concentration  Range:  0.3-5 mg/L using  a  wavelength of  232.0 run
Sensitivity:   0.15  mg/L
Approximate Detection  Limit:  0.04 mg/L

Preparation of Standard  Solution

1.   Stock Solution:  Dissolve 4.953 g of nickel nitrate, Ni(N03)2'6H20
     (analytical reagent grade) in deionizing distilled water.  Add 10 mL of
     cone, nitric acid and dilute to 1 liter deionized distilled water.  1
     mL  - 1 mg Ni (1000 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  The calibration standards should be
     prepared using the same type of acid and at the same concentration as
     will result in the sample to be analyzed after sample preparation.

Instrumental Parameters  (General)

      1.  Nickel hollow cathode lamp
      2.  Wavelength:   232,0 nm
      3.  Fuel:  Acetylene
      4.  Oxidant:   Air
      5.  Type of flame:   Oxidizing
Interferences

1.   The 352.4 nm wavelength is less susceptible to spectral interference
     and may be used.  The calibration curve is more linear at this
     wavelength; however, there is some loss of sensitivity.

Notes

1.   For concentrations of nickel below 100 ug/L,  use of the  Furnace
     Technique (Method 249.2 CLP-M) is recommended.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-123                               ILM01.0

-------
                                    SILVER*

          Method 272.1 CLP-M**  (Atomic Absorption, Flame Technique)


 Optimum Concentration Range:   0.1-4 mg/L using a wavelength of 328.1 nm
 Sensitivity:   0.06 mg/L
 Approximate Detection Limit:   0.01 mg/L

 Preparation of Standard Solution

 1.    Stock  Solution:   Dissolve  1.575 g  of AgNO^,  (analytical reagent grade)
      in deionized  distilled water,  add  10 mL cone. HNO-j and make up  to  1
      liter.   1 mL  - 1 mg Ag (1000  mg/L).

 2.    Prepare  dilutions of  the stock solution to be used as calibration
      standards at  the time of analysis.   The calibration standards must be
      prepared using the  same type  of acid and  at  the same concentration as
      will result in the  sample  to  be analyzed  after sample preparation.

 3.    Iodine Solution,  1  N:  Dissolve 20 grams  of  potassium iodide, KI
      (analytical reagent grade)  in 50 mL  of deionized distilled water, add
      12.7 grains of iodine, 1^  (analytical reagent grade) and dilute  to 100
      mL.  Store in a  brown bottle.

 4.    Cyanogen Iodide  (CNI) Solution:  To  50 mL of deionized distilled water
      add 4.0  mL cone.  NH^OH, 6.5  grams KCN, and  5.0 mL of 1.0 N I2
      solution.  Mix and  dilute  to  100 mL  with  deionized distilled water.
      Fresh solution should be prepared every two weeks.(1)

 Instrumental  Parameters  (General)

       1.  Silver hollow  cathode  lamp
       2.  Wavelength:   328.1 nm
       3.  Fuel:  Acetylene
       4. . Oxidant:   Air
       5.  Type of flame:  Oxidizing
 Notes

 1.   For concentrations  of silver below 30 ug/L, use of the Furnace
     Technique (Method 272.2 CLP-M)  is recommended.

 2.   Silver nitrate standards are light sensitive.  Dilutions of the stock
     should be discarded after use as concentrations below 10 mg/L are not
     stable over long periods of time.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-124                               ILM01.0

-------
3.   If absorption to container walls or the formation of AgCl is suspected,
     make the sample basic using cone. NH^OH and add 1 mL of (CNI) solution
     per 100 mL of sample.  Mix the sample and allow to stand for 1 hour
     before proceeding with the analysis.(1)

4.   The 338.2 run wavelengrh may also be used.   This has a relative
     sensitivity of 2.

References

1.   "The Use of Cyanogen Iodide (CNI) as a Stabilizing Agent for Silver in
     Photographic Processing Effluent Sample",  Owerbach,  Daniel,
     Photographic Technology Division, Eastman Kodak Company,  Rochester,
     N.Y.  14650.
                                  D-125                                ILM01.0

-------
                                  THALLIUM*

          Method 279.1  CLP-M**  (Atomic  Absorption,  Flame Technique)


Optimum Concentration Range:  1-20 mg/L using a wavelength of 276.8 nm
Sensitivity:  0.5 mg/L
Approximate Detection Limit:  0.1 mg/L

Preparation of  Standard Solution

1.   Stock Solution:  Dissolve 1.303 g of thallium nitrate,  T1N03
     (analytical reagent grade) in deionized distilled water.   Add 10 mL of
     cone,  nitric acid and dilute to 1 liter with deionized distilled
     water.  1 mL - 1 rag Tl (1000 mg/L).

2.   Prepare dilutions of the stock solution to be used as calibration
     standards at the time of analysis.  The calibration standards must be
     prepared using nitric acid and at the same concentration as will result
     in the sample to be analyzed after sample preparation.

Instrumental Parameters (General)

      1.  Thallium hollow cathode lamp
      2.  Wavelength:   276.8 nm
      3.  Fuel:   Acetylene
      4.  Oxidant:   Air
      5.  Type of flame:   Oxidizing

      Notes

1.   For concentrations of thallium below 0.2 mg/L,  use of the Furnace
     Technique  (Method 279.2 CLP-M) is recommended.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-126                                ILM01.0

-------
                                  VANADIUM*

          Method 286.1 CLP-M** (Atomic Absorption, Flame Technique)


Optimum  Concentration Range:   2-100 mg/L using  a  wavelength  of  318.4  run
Sensitivity:  0.8 mg/L
Approximate Detection Limit:   0.2 mg/L

Preparation of  Standard  Solution

1.   Stock Solution:  Dissolve 1.7854 g of vanadium pentoxide, V20c
     (analytical reagent grade) in 10 mL of cone, nitric acid and dilute to
     1 liter with deionized distilled water.  1 mL — 1 mg V  (1000 mg/L).

2.   Aluminum nitrate solution:  Dissolve 139 g aluminum nitrate,
     Al(N03)3'9H20,  in 150 mL of deionized distilled water; heat to effect
     solution.  Allow to cool and make up to 200 mL.

3.   Prepare dilutions of the stock vandium solution to be used as
     calibration standards at the time of analysis.  The calibration
     standards must be prepared using the same type of acid and at the same
     concentration as will result in the sample to be analyzed after sample
     preparation.  To each 100 mL of standard and sample alike,  add 2 mL of
     the aluminum nitrate solution.

Insjtrumental Parameters  (General')

      1.  Vanadium hollow cathode lamp
      2.  Wavelength:   318.4  run
      3.  Fuel:   Acetylene
      4.  Oxidant:   Nitrous  Oxide
      5.  Type of flame:   Fuel rich

Interferences

1.   It has been reported chat high concentrations of aluminum and titanium
     increase the sensitivity of vanadium.   This interference can be
     controlled by adding excess aluminum (1000 ppm) to both samples and
     standards.   [Talanta 15, 871 (1968)].

Notes

1.   For concentrations of vanadium below 0.5 mg/L, use of the Furnace
     Technique (Method 282.6 CLP-M)  is recommended.
*This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.
                                                                       ILM01.0

-------
                                    ZINC*

          Method 289.1 CLP-M** (Atomic Absorption, Flame Technique)


 Optimum Concentration Range:   0.05-1  rog/L using a wavelength of 213.9 nm
 Sensitivity:   0.02  mg/L
 Approximate Detection Limit:   0.005 mg/L

 Preparation of Standard Solution

 1.    Stock  Solution:   Carefully  weigh 1.00  g of zinc metal  (analytical
      reagent  grade)  and  dissolve cautiously in  10 mL HNO^.  When solution  is
      complete make up  to 1  liter with deionized distilled water.   1 mL - 1
      rag Zn  (1000 mg/L).

 2.    Prepare  dilutions of the stock solution to be used as calibration
      standards at the  time  of analysis.  The calibration standards must be
      prepared using  the  same type of acid and at the same concentration as
      will result in  the  sample to be analyzed after sample preparation.

 Instrumental  Parameters  (General)

       1.  Zinc hollow cathode lamp
       2.  Wavelength:  213.9 nm
       3.  Fuel:  Acetylene
       4.  Oxidant:   Air
       5.  Type of flame:  Oxidizing

Notes

1.   High levels of silicon may interfere.

2.   The air-acetylene flame absorbs about 25% of the energy at the 213.9 nra
     line.

3.   The sensitivity may be  increased by the use of low-temperature flames.

A.   Some container cap liners  can  be  a source of zinc contamination.   To
     circumvent or  avoid this problem,  the use of the polypropylene caps  is
     recommended.

5.   For concentrations of zinc below  0.01 mg/L, use  of the  Furnace
     Technique (Method 289.2 CLP-M)  is  recommended.
-'This method may only be used under specified conditions.

**CLP-M Modified for the Contract Laboratory Program.

                                  D-128                               ILM01.0

-------
                              EXHIBIT E







           QUALITY ASoURANCE/QUALITY CONTROL REQUIREMENTS







                                                           Page  No.




SECTION I   - GENERAL QA/QC PRACTICES                        E-l




SECTION II  - SPECIFIC QA/QC PROCEDURES                      E-2




SECTION III - QUALITY ASSURANCE PLAN                         E-5




SECTION IV  - DATA MANAGEMENT                                E-7




SECTION V   - REQUIRED QA/QC OPERATIONS                      E-9




SECTION VI  - LABORATORY EVALUATION PROCESS                  E-25
                                                                  ILH01.0

-------
                                 SECTION I

                          GENERAL QA/QC PRACTICES
Standard laboratory practices for laboratory cleanliness as applied to
glassware and apparatus must be adhered to.  Laboratory practices with
regard to reagents, solvents, and gases must also be adhered to.  For
additional guidelines regarding these general laboratory procedures, see
Sections 4 and 5 of the Handbook for Analytical Quality Control in Water
and Wastewater Laboratories EPA-600/4-79-019,  USEPA Environmental
Monitoring and Support Laboratory, Cincinnati, Ohio, March 1979.
                                    E-2                                ILM01.0

-------
                                 SECTION II

                          SPECIFIC  QA/QC PROCEDURES


The quality assurance/quality control (QA/QC) procedures defined herein
must be used by the Contractor when performing the methods specified in
Exhibit D.  When additional QA/QC procedures are specified in the methods
in Exhibit D, the Contractor must also  follow these procedures.  NOTE:  The
cost of performing all QA/QC procedures specified in this Statement of Work
is included in the price of performing  the bid lot,  except for duplicate,
spike, and laboratory control sample analyses, which shall be considered
separate sample analyses.

The purpose of this document is to provide a uniform set of procedures for
the analysis of inorganic constituents of samples, documentation of methods
and their performance, and verification of the sample data generated.  The
program will also assist laboratory personnel in recalling and defending
their actions under cross examination if required to present court
testimony in enforcement case litigation.

The primary function of the QA/QC program is the definition of procedures
for the evaluation and documentation of sampling and analytical
methodologies and the reduction and reporting of data.   The objective is to
provide a uniform basis for sample collection and handling, instrument and
methods maintenance,  performance evaluation, and analytical data gathering
and reporting.  Although it is impossible to address all analytical
situations in one document, the approach taken here is to define minimum
requirements for all  major steps relevant to any inorganic analysis.  In
many instances where  methodologies are available, specific quality control
procedures are incorporated into the method documentation (Exhibit D).
Ideally, samples involved in enforcement actions are analyzed only after
the methods have met  the minimum performance and documentation requirements
described in this document.

The Contractor is required to participate in the Laboratory Audit and
Intercomparison Study Program run by EPA EMSL-Las Vegas.  The Contractor
can expect to analyze at least two samples per calendar quarter during the
contract period.

The Contractor must perform and report to SMO and EMSL as specified in
Exhibit B quarterly verification of instrument detection limits (IDL) by
the method specified  in Exhibit E,  by type and model for each instrument
used on this contract.  All the IDLs must meet the CRDLs specified in
Exhibit C.   For ICP methods,  the Contractor must also report,  as specified
in Exhibit B,  linearity range verification, all interelement correction
factors, wavelengths  used,  and integration times.

In this Exhibit,  as well as other places within this Statement of Work,  the
term "analytical  sample" is used in discussing the required frequency or
placement of certain  QA/QC measurements.  The term "analytical sample" is
defined in the glossary, Exhibit G.   As  the term is  used, analytical sample
includes all field samples, including Performance Evaluation samples,
received from an external source,  but it also includes  all required QA/QC
samples (matrix spikes,  analytical/post-digestion spikes, duplicates,

                                   E-3                                 ILMOL.O

-------
serial dilutions, LCS,  ICS, CRDL standards, preparation blanks and linear
range analyses) except those directly related to instrument calibration or
calibration verification (calibration standards, ICV/ICB, CCV/CCB) .   A
"frequency of 10%" means once every 10 analytical samples.  Note:
Calibration verification samples (ICV/CCV) and calibration verification
blanks (ICB/CCB) are not counted as analytical samples when determining 10%
frequency.

In order for the QA/QC information to reflect the status of the samples
analyzed, all samples and their QA/QC analysis must be analyzed under the
same operating and procedural conditions.

If any QC measurement fails to meet contract criteria, the analytical
measurement may not be repeated  prior to taking the appropriate corrective
action as specified in Exhibit E.

The Contractor must report all QC data in the exact format specified in
Exhibits B and H.

Sensitivity,  instrumental detection limits (IDL's),  precision, linear
dynamic range and interference effects must be established for each analyte
on a particular instrument.  All reported measurements must be within the
instrumental linear ranges.  The analyst must maintain quality control data
confirming instrument performance and analytical results.

In addition,  the Contractor shall establish a quality assurance program
with the objective of providing sound analytical chemical measurements.
This program shall incorporate the quality control  procedures, any
necessary corrective action,  and all documentation  required during data
collection as well as the quality assessment measures performed by
management to ensure acceptable data production.

As evidence of such a program,  the Contractor shall  prepare a written
Quality Assurance Plan (QAP)  (see Section III) which describes the
procedures that are implemented to achieve the following:

      Maintain data integrity,  validity,  and useability.

    • Ensure  that analytical measurement systems are maintained in an
      acceptable state  of stability and reproducibility.

      Detect  problems through data assessment and establishes corrective
      action procedures which keep the analytical process reliable.

      Document all aspects of the measurement process in order to provide
      data which are technically sound and legally  defensible.
                                   E-4                                 ILM01.0

-------
                                 SECTION III

                           QUALITY ASSURANCE PLAN

The QAP must present, in specific terms,  the policies, organization,
objectives, functional guidelines, and specific QA and QC activities
designed to achieve the data quality requirements in this contract.  Where
applicable, SOPs pertaining to each element shall be included or referenced
as part of the QAP.  The QAP must be available during on-site laboratory
evaluation and upon written request by the APO.  The elements of the QAP
are listed in the following outline.

A.   Organization and Personnel

     1.    QA  Policy and  Objectives

     2.    QA  Management

          a.   Organization

          b.   Assignment of QC and QA Responsibilities

          c.   Reporting Relationships

          d.   QA Document Control Procedures

          e.   QA Program Assessment Procedures

     3.    Personnel

          a.   Resumes

          b.   Education and Experience Pertinent  to  this Contract

          c.   Training Progress

B.   Facilities  and Equipment

     1.    Instrumentation and  Backup Alternatives

     2.    Maintenance Activities  and Schedules

C.   Document Control

     1.    Laboratory Notebook  Policy

     2.    Samples Tracking/Custody Procedures

     3.    Logbook Maintenance  and Archiving Procedures

     4.    SDG File  Organization,  Preparation and Review  Procedures

     5.    Procedures for Preparation, Approval, Review,  Revision, and
          Distribution of SOPs


                                   E-5                                 ILM01.0

-------
     6.   Process for Revision of Technical or Documentation Procedures

D.   Analytical Methodology

     1.   Calibration Procedures and Frequency

     2.   Sample Preparation Procedures

     3.   Sample Analysis Procedures

     4.   Standards Preparation Procedures

     5.   Decision Processes,  Procedures,  and Responsibility for Initiation
          of Corrective Action

E.   Data Generation

     1.   Data Collection Procedures

     2.   Data Reduction Procedures

     3.   Data Validation Procedures

     4.   Data Reporting and Authorization Procedures

F.   Quality Assurance

     1.   Data Quality Assurance

     2.   Systems/Internal Audits

     3.   Performance/External Audits

     4.   Corrective Action Procedures

     5.   Quality Assurance Reporting Procedures

     6.   Responsibility Designation

G.   Quality Control

     1.   Solvent,  Reagent and Adsorbent Check Analysis

     2.   Reference Material Analysis

     3.   Internal Quality Control Checks

     4.   Corrective Action and Determination of  QC  Limit  Procedures

     5.   Responsibility Designation
                                    E-6                                 ILM01.0

-------
                                  SECTION IV

                               DATA MANAGEMENT

Data management procedures  are  defined as procedures specifying the
acquisition or entry, update, correction, deletion, storage and security of
computer readable  data and  files.  These procedures should be in written
form and contain a clear definition  for all databases and files used to
generate or resubmit deliverables.   Key areas of concern include:  system
organization  (including personnel  and security), documentation operations,
traceability  and quality control.

Data manually entered from  hard-copy must be quality controlled and the
error rates estimated.  Systems should prevent entry of incorrect or out-
of-range data and  alert data entry personnel of errors.  In addition, data
entry error rates  must be estimated  and recorded on a monthly basis by
reentering a  statistical sample of the data entered and calculating
discrepancy rates  by data element.

The record of changes in the form  of corrections and updates to data
originally generated, submitted, and/or resubmitted must be documented no
allow traceablilty of updates.  Documentation must include the following
for each change:

o   Justification or rationale for the change.
o   Initials  of the person making the change  or  changes.   Data  changes  must
    be  implemented and reviewed by a person or group  independent  of the
    source  generating the  deliverable.
o   Change  documentation must be retained  according to  the  schedule of  the
    original  deliverable.

c   Resubmitted diskettes  or other deliverables  must  be  reinspected as  a
    part of the laboratories'  internal inspection process  prior  to
    resubmission.   The  entire deliverable, not just the  changes,  must be
    inspected.

o   The  Laboratory Manager must  approve  changes  to  originally submitted
    deliverables.

o   Documentation of data  changes  may be requested  by laboratory  auditors.
LIfecycle management procedures must: be applied Co  computer software
systems developed by the laboratory  to be used to generate and edit
contract deliverables.   Such systems must be  thoroughly tested and
documented prior to utilization.

o   A software test and acceptance plan  including test  requirements,  test
    results and acceptance  criteria must be developed,  followed, and
    available  in  written form.

o   System  changes  must not be made directly  to  production  systems
    generating deliverables.  Changes must be  made  first  to  a development
    system  and tested prior to  implementation.

o   Each version  of the production system will be given  an  identification
    number, date  of installation,  date of  last operation and archived.
                                    E-7                                 ILM01  0

-------
o   System and operations  documentation must be  developed  and  maintained for
    each system.   Documentation must  include a users  manual  and  an
    operations and maintenance manual.
Individual(s) responsible for the following functions must be identified:

o   System operation and maintenance  including documentation and training.

o   Database  integrity, including data entry, data  updating  and  quality
    control.

o   Data and  system security, backup  and archiving.
                                   E-8                                 ILM01.0

-------
                                 SECTION V

                         REQUIRED QA/QC OPERATIONS

This section outlines the minimum QA/QC operations necessary to satisfy the
analytical requirements of the contract.   The following QA/QC operations
must be performed as described in this Exhibit:

     1.    Instrument Calibration

     2.    Initial  Calibration Verification  (ICV)  and  Continuing Calibration
          Verification (CCV)

     3.    CRDL  Standards  for  AA (CRA)  and ICP (CRI)
     4.    Initial  Calibration Blank  (ICB),  Continuing Calibration  Blank
          (CCB), and Preparation Blank (PB) Analyses

     5.    ICP Interference  Check Sample (ICS) Analyses
     6.    Spike Sample Analysis (S)
     7.    Duplicate  Sample  Analysis  (D)

     8.    Laboratory Control  Sample  (LCS) Analysis
     9.    ICP Serial Dilution Analysis (L)

     10.   Instrument Detection Limit (IDL)  Determination

     11.   Interelement Corrections for ICP  (ICP)
     12.   Linear Range Analysis (LRA)

     13.   Furnace  AA QC Analyses


- •    Instrument Calibration

     Guidelines for  instrumental calibration  are  given in EPA 600/4-79-020
     and/or Exhibit  D. Instruments must be  calibrated daily or once every 24
     hours and  each  time  the  instrument is  set up.   The instrument
     standardization date and time must be  included in the raw data.

     For atomic absorption systems,  calibration standards are prepared by
     diluting the  stock metal solutions at  the time of analysis.  Date and
     time of preparation  and  analysis must  be given in the raw data.

     Calibration  standards  must be prepared fresh each time an analysis is
     to be made and  discarded after  use.   Prepare a blank and at least three
     calibration  standards  in graduated amounts in the appropriate range.
     One atomic absorption calibration standard must be at the CRDL except
     for mercury.  The calibration standards  must be prepared using the same
     type of acid  or combination of  acids  and at the same concentration as
     will result  in  the samples following  sample preparation.

     Beginning  with  the blank, aspirate or  inject the standards and record
     the readings. If the AA  instrument configuration prevents the required
     4-point calibration,  calibrate  according to instrument manufacturer's
     recommendations, and analyze the remaining required standards
     immediately  after calibration.   Results  for these standards must be
     within + 5%  of  the true  value.   Each  standards concentration and the
                                    E-9                                ILM01.0

-------
     calculations to show that the +5% criterion has been met,  must be given
     in the raw data.  If the values do not fall within this  range,
     recalibration is necessary.

     The + 5% criteria does not apply to the atomic absorption  calibration
     standard at the CRDL.

     Calibration standards  for AA procedures must be  prepared  as  described
     in Exhibit D.

     Baseline correction is acceptable as long as it is performed  after
     every sample or after  the continuing calibration verification and blank
     check; resloping is acceptable as long as it is immediately preceded
     and immediately followed by  CCV and CCB.   For cyanide and  mercury,
     follow the calibration procedures outlined in  Exhibit D.   One cyanide
     calibration standard must be at the CRDL.   For ICP systems, calibrate
     the instrument according to  instrument manufacturer's recommended
     procedures.   At least  two standards must be used for ICP calibration.
     One of the standards must be a blank.

2.    Initial Calibration Verification CICV)  and Continuing Calibration
     Verification (CCV)

     a.    Initial Calibration Verification  (ICV)

          Immediately after each  of the ICP,  AA and cyanide systems  have
          been calibrated,  the accuracy of  the  initial calibration shall  be
          verified  and documented for every  analyte by the analysis  of EPA
          Initial Calibration Verification  Solution(s)  at each  wavelength
          used for  analysis.   When measurements exceed the control limits of
          Table 1-Initial and Continuing Calibration Verification  Control
          Limits  for Inorganic Analyses (in  Exhibit E),  the analysis must be
          terminated,  the problem corrected,  the  instrument recalibrated,
          and the calibration reverified.

          If the  Initial Calibration Verification Solution(s) are  not
        .  available from EPA,  or  where a certified solution of  an  analyte is
          not available from any  source,  analyses shall  be conducted on an
          independent standard at a concentration other  than  that  used for
          instrument calibration,  but within the  calibration  range.   An
          independent standard is defined as a  standard  composed of  the
          analytes  from a different source  than those used in the  standards
          for the instrument calibration.

          For ICP,  the Initial Calibration Verification  Solution(s)  must  be
          run at  each wavelength  used for analysis.   For CN,  the initial
          calibration verification  standard must be distilled.  The Initial
          Calibration Verification for CN serves  as a Laboratory Control
          Sample; thus it must be distilled with the batch of samples
          analyzed in association with that  ICV.   This means  that  an ICV
          must be distilled with  each batch  of samples analyzed and that  the
          samples distilled with  an ICV must be analyzed with that
          particular ICV.  The values for the  initial and subsequent
          continuing calibration  verifications  shall be  recorded on FORM  II-
          IN for ICP, AA, and cyanide analyses,  as indicated.

                                    E-10                                ILM01.0

-------
b.   Continuing Calibration Verification (CCV)

     To ensure calibration accuracy during each analysis run,  one of
     the following standards is to be used for continuing calibration
     verification and must be be analyzed and reported for every
     wavelength used for the analysis of each analyte, at a frequency
     of 10% or every 2 hours during an analysis run,  whichever is tr.ore
     frequent.   The standard must also be analyzed and reported for
     every wavelength used for analysis at the beginning of the run and
     after the last analytical sample.  The analyte concentrations in
     the continuing calibration standard must be one  of the following
     solutions at or near the mid-range levels of the calibration
     curve:
          1.  EPA Solutions
          2.  NBS SRM 1643a
          3.  A Contractor-prepared standard solution
         TABLE 1.  INITIAL AND CONTINUING CALIBRATION VERIFICATION
                   CONTROL LIMITS FOR INORGANIC ANALYSES
Analytical Method
ICP/AA
Cold Vapor AA
Other
Inorganic
Species
Metals
Mercury
Cyanide
% of True
Low Limit
90
80
85
Value (EPA Set)
High Limit
110
120
115
    The same  continuing calibration  standard must be  used  throughout
    the analysis  runs  for a Case  of  samples  received.

    Each CCV  analyzed  must reflect the  conditions of  analysis of all
    associated  analytical samples (the  preceding 10 analytical samples
    or the preceding analytical samples up to the previous CCV),  The
    duration  of analysis,  rinses  and other related operations that may
    affect the  CCV measured result may  not be applied to the CCV to a
    greater extent than the extent applied to the associated
    analytical  samples.   For instance,  the difference in time between
    a CCV analysis and the blank  immediately following it as well as
    the difference in  time between the  CCV and the analytical sample
    immediately preceding it may  not exceed  the lowest difference in
    time between  any two consecutive analytical samples associated
    with the  CCV.

    If the deviation of the continuing  calibration verification is
    greater than  the control limits  specified in Table 1-Initial and
    Continuing  Calibration Verification Control Limits for Inorganic
    Analyses, the analysis must be stopped,  the problem corrected, the
    instrument  must be  recalibrated, the calibration  verified and the
    reanalysis  of preceding 10 analytical samples or  all analytical

                              E-ll                               ILM01.0

-------
          samples  analyzed since the last compliant calibration verification
          must be  performed for the analytes affected.  Information
          regarding  the  continuing verification of calibration shall be
          recorded on  FORM II-IN for ICP, AA and cyanide as indicated.

3.    CRDL Standards  for  ICP  (CRI) and A A (CRA)

     To verify  linearity near  the CRDL for  ICP analysis, the  Contractor  must
     analyze an  ICP  standard  (CRI) at  two times the CRDL or two  times  the
     IDL,  whichever  is greater, at the beginning and  end of each  sample
     analysis run, or  a  minimum of twice per 8 hour working shift, whichever
     is more frequent, but not before  Initial Calibration Verification.
     This standard must  be run by ICP  for every wavelength used  for
     analysis, except  those  for Al, Ba, Ca, Fe, Mg, Na  and K.

     To verify linearity near  the CRDL for  AA analysis, the Contractor must-
     analyze an  AA standard  (CRA) at the CRDL or the  IDL, whichever  is
     greater, at the beginning of each sample analysis  run, but not  before
     the Initial Calibration Verification.

     Specific acceptance criteria for  the two standards will  be set  by EPA
     in the future.  In  the  interim, the Contractor must analyze  and report
     these Standards on  FORM II(PART 2)-IN.

4.    Initial Calibration Blank (ICB).  Continuing Calibration  Blank  (CCB).
     and Preparation Blank (PB) Analyses

     a.   Initial  Calibration  Blank (ICB) and Continuing Calibration Blank
          (CCB)  Analyses

          A calibration  blank  must be  analyzed at each wavelength used for
          analysis immediately after every  initial and continuing
          calibration  verification, at a frequency of 10% or  every 2 hours
          during the run, whichever is more frequent.  The blank  must  be
          analyzed at  the beginning of the  run and after the  last
          analytical sample.   Note:  A CCB must be run after  the  last  CCV
          that was run after the last  analytical sample of the run.  The
          results  for  the calibration  blanks shall be recorded on FORM III-
          IN for ICP,  AA and cyanide analyses, as indicated.  If  the
          magnitude  (absolute  value) of the calibration blank result exceeds
          the IDL, the result  must be  so reported in ug/L on  FORM III-IN,
          otherwise  report as  IDL-U.   If the absolute value blank result
          exceeds  the  CRDL (Exhibit C), terminate analysis, correct  the
          problem, recalibrate, verify the  calibration and reanalyze the
          preceding  10 analytical samples or all analytical samples  analyzed
          since  the  last compliant calibration blank.

     b.   Preparation  Blank  (PB) Analysis

          At least one preparation blank (or reagent blank),  consisting  of
          deionized  distilled  water processed through each sample
          preparation  and analysis procedure (See Exhibit D,  Section III),
                                   E-12                                ILM01.0

-------
         must  be  prepared and analyzed with every Sample Delivery Group,  or
         with  each batch  of samples digested, whichever is more frequent.

         The first batch of samples in an SDG is to be assigned to
         preparation  blank one, the second batch of samples to preparation
         blank two, etc. (see FORM 1II-IN).  Each data package must contain
         the results  of all the preparation blank analyses associated with
         the samples  in that SDG.

         This  blank is to be reported for each SDG and used in all analyses
         to ascertain whether sample concentrations reflect contamination
         in the following manner:

         1)    If  the  absolute value of the concentration of the blank is
               less than or equal to the Contract Required Detection Limit
               (Exhibit C),  no correction of sample results is performed.

         2)    If  any  analyte concentration in the blank is above the CRDL,
               the lowest concentration of that analyte in the associated
               samples must be lOx the blank concentration.   Otherwise, all
               samples associated with che blank with the analyte's
               concentration less than lOx the blank concentration and above
               the CRDL, must be redigested and reanalyzed for that analyte
               (except for an identified aqueous soil field blank).   The
               sample  concentration is not to be corrected for the blank
               value.

         3)    If  the  concentration of the blank is below the negative CRDL,
               then all samples reported below lOx CRDL associated with the
               blank must be redigested and reanalyzed.

         The values for the preparation blank must be recorded in ug/L for
         aqueous  samples and in isg/Kg for solid samples on FORM III-IN for
         ICP,  AA, and cyanide analyses.

5.    ICP Interference  Check  Sample  (ICS) Analysis

     To verify  interelement  and background correction  factors,  the
     Contractor must analyze and report the  results  for  the ICP Interference
     Check  Samples at  the beginning and end  of  each  analysis run or a
     minimum of twice  per 8 hour working shift, whichever is more frequent,
     but not before  Initial  Calibration Verification.  The ICP  Interference
     Check  Samples must be obtained from EPA (EMSL/LV) if available and
     analyzed  according to the instructions  supplied with the ICS.

     The Interference  Check  Samples consist  of  two solutions: Solution A  and
     Solution AB.   Solution  A consists of the  interferents, and Solution  AB
     consists  of the analytes mixed with the interferents.  An  ICS analysis
     consists  of analyzing both  solutions consecutively  (starting with
     Solution A) for all wavelengths  used for  each analyte reported by ICP.
 A group of samples prepared at Che same time.
                                   E-13                                ILM01.0

-------
Results for the ICP analyses of Solution AB during the analytical runs
must fall within the control limit of +20% of the true value for the
analytes included in the Interference Check Samples.  If not, terminate
the analysis,  correct the problem, recalibrate the instrument, and
reanalyze the analytical samples analyzed since the last good ICS.  If
true values for analytes contained in the ICS and analyzed by ICP are
not supplied with the ICS,  the mean must be determined by initially
analyzing the ICS at least five times repetitively for the particular
analytes.  This mean determination must be made during an analytical run
where the results for the previously supplied EPA ICS met all contract
specifications.  Additionally, the result of this initial mean
determination is to be used as the true value for the lifetime of that
solution (i.e., until the solution is exhausted).

If the ICP Interference Check Sample is not available from EPA,
independent ICP Check Samples must be prepared with interferent and
analyte concentrations at the levels specified in Table 2-Interferent
and Analyte Elemental Concentrations Used for ICP Interference Check
Sample.  The mean value and standard deviation must be established by
initially analyzing the Check Samples at least five ^imes repetitively
for each parameter on FORM IV-IN.   Results must fall within the control
limit of +20%  of the established mean value.   The mean and standard
deviation must be reported in the raw data.   Results from the
Interference Check Sample analyses must be recorded on FORM IV-IN for
all ICP parameters.
TABLE 2.  INTERFERENT AND ANALYTE ELEMENTAL CONCENTRATIONS USED FOR ICP
                       INTERFERENCE CHECK SAMPLE
  Analytes
(mg/L)
Interferents
(mg/L)
AS
Ba
Be
- Cd
Co
Cr
Cu
Mn
Ni
Pb
V
Zn
1
0
0
1
0
0
0
0
1
1
0
1
.0
.5
.5
.0
.5
.5
.5
.5
.0
.0
.5
.0
Al
Ca
Fe
Mg








500
500
200
500








Spike Sample Analysis (S)

The spike sample analysis is designed to provide information about the
effect of the sample matrix on the digestion and measurement
methodology.  The spike is added before the digestion (i.e., prior to
the addition of other reagents) and prior to any distillation steps
(i.e., CN-).  At least one spike sample analysis must be performed on
                              E-14
                                               ILM01.0

-------
      each  group of  samples  of a similar matrix  type  (i.e., water,  soil) and
      concentration  (i.e., low, medium) or for each Sample Delivery Group.

      If  the  spike analysis  is performed on the  same  sample that  is chosen
      for the duplicate sample analysis, spike calculations must  be performed
      using the results of the sample designated as the "original sample"
      (see  section 7,  Duplicate Sample Analysis).  The average of the
      duplicate results cannot be used for the purpose of determining percent
      recovery.  Samples  identified as field blanks cannot be used  for spiked
      sample  analysis.  EPA  may require that a specific sample be used for
      the spike sample analysis.

      The analyte spike must be added in the amount given .in Table  3-Spiking
      Levels  for Spike Sample Analysis, for each element analyzed.  If two
      analytical methods  are used to obtain the  reported values for the same
      element within a Sample Delivery Group (i.e. ICP, GFAA),  spike samples
      must  be run by each method used.

      If  the  spike recovery  is not at or within  the limits of 75-125%, the
      d->ta  of all samples received associated with that spike sample and
      determined by  the same analytical method must be flagged with the
      letter  "N" on FORMs I-IN and V-IN.  An exception to this rule is
      granted in situations  where the sample concentration exceeds  the spike
      concentration by a  factor of four or more.   In  such an event, the data
      shall be reported unflagged even if the percent recovery does not meet
      the 75-125% recovery criteria.

      For flame AA, ICP,  and CN analyses, when the pre-digestion/pre-
      distillation spike  recovery falls outside  the control limits  and the
      sample  result does not exceed 4x the spike added, a post-
      digestion/post-distillation spike must be performed for those elements
      that do not meet the specified criteria (exception:  Ag).   Spike the
      unspiked aliquot of the sample at 2x the indigenous level or  2x CRDL,
     whichever is greater.  Results of the post-digestion/post-distillation
      spike must be reported on FORM V(PART 2)-IN.  Note:   No pose  digest
      spike is required for  Hg.

      In the  instance where  there is more thaa one spike sample per matrix
      and concentration per  method per SDG, if one spike sample recovery is
     not within contract criteria,  flag all the samples of the same matrix,
      level, and method in the SDG.   Individual component percent recoveries
      (%R) are calculated as follows:

                         %Recovery  = (SSR-SR) x 100
                                         SA
      Where,    SSR  =  Spiked Sample Result
                SR  =  Sample Result
                SA  =  Spike Added

     When sample  concentration is less than the instrument detection limit,
     use  SR = 0 only for purposes of calculating % Recovery.   The spike
2
 EPA may require additional spike sample analysis, upon Administrative
 Project Officer request, for which  the Contractor will be paid.

                                   E-15                                ILM01.

-------
sample results, sample results and % Recovery (positive or negative)
must be reported on FORM V-IN for ICP, AA and cyanide analyses, as
indicated.

The units for reporting spike sample results will be-identical to those
used for reporting sample results in FORM I-IN (i.e.,  ug/L for aqueous
and mg/Kg dry weight basis for solid).
                              E-16                                ILM01.0

-------
           TABLE 3.   SPIKING LEVELS  FOR SPIKE  SAMPLE ANALYSIS
                    For ICP/AA
For Furnace AA
Other
                                                                  (1)
Element

Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide
Water
(ug/L)
2,000
500
2,000
2,000
50
50
*
200
500
250
1,000
500
*
500
500
*
2,000
50
*
2,000
500
500

Soil(2) Water Soil(2)
(mg/kg) (ug/L) (rag/kg)
•x
100 100 20
400 40 8
400
10
10 5 1
*
40
100
50
*
100 20 4
*
100
1
100
*
400 10 2
10
*
400 50 10
100
100
10°(3)
 No spike required.  NOTE:  Elements without spike levels and not
designated with an asterisk, must be spiked at appropriate levels.

 Spiking level reported is for both water and soil/sediment matrices.
2
 The levels shown indicate concentrations in the final digestate of the
spiked sample (100 mL for mercury and 200 mL for all other metals) when
the wet weight of 1 gram (for ICP, Furnace, and Flame AA), or 0.2 grams
(for mercury) of sample is taken for analysis.  Adjustment must be made
to maintain these spiking levels when the weight of sample taken
deviates by more than 10% of these values.  Appropriate adjustment must
be made for microwave digestion procedure where 0.5 grams of sample or
50.0 mL (45.0 mL of sample plus 5.0 mL of acid) of aqueous sample are
required for analysis.

 The level shown indicates the amount of cyanide that must be added to
the origin.il (undistilled) sample.  For instance, 100 ug must be added
per each Liter of aqueous sample.  If the sample volume is 500 mL, then
50 ug of cyanide must be added.  If the volume is 50 mL, then 5 ug of
cyanide must be added.
For soil samples, 25 ug of cyanide must be added per each gram of solid
sample taken for analysis.  The spiking level is dependent on the
weight of the sample taken and the final distillate volume.  If one
gram of sample is taken for analysis, and the final distillate volume
is 250 mL, then the distillate must contain cyanide at a concentration
                              E-17
                         ILM01.0

-------
     of 100 ug/L.  If five grams of sample are taken,  then the distillate
     must contain cyanide at a concentration of 500 ug/L.   Assuming a sample
     of one gram, the manual and semi-automated colorimetric methods call
     for a cyanide concentration of 25 ug per the 500  mL mixture of the
     sample, reagents, and water before distillation.   The final distillate,
     in this case, contains cyanide at a concentration of 100 ug/L.  For the
     midi-distillation method, a cyanide concentration of 25 ug must be
     added into the 50 mL mixture of sample, reagents, and water before
     distillation.  This yields a cyanide concentration of 500 ug/L in the
     final distillate of 50 mL.
7.   Duplicate Sample Analysis CD)

     One duplicate sample must be analyzed from each group of samples of a
     similar matrix type (i.e.,  water,  soil)  and concentration (i.e., low,
     medium) or for each Sample Delivery Group.    Duplicates cannot be
     averaged for reporting on FORM I-IN.

     Duplicate sample analyses are  required for  percent  solids.  Samples
     identified as field blanks cannot  be  used for duplicate sample
     analysis.  EPA may require that a  specific  sample be  used for duplicate
     sample analysis.  If two analytical methods are used  to obtain the
     reported values for the same element  for a  Sample Delivery  Group (i.e.,
     ICP, GFAA),  duplicate samples  must be run by each method used.

     The relative percent differences (RPD) for  each component are
     calculated as follows:

                       RPD  - IS -  PI x 100
                              (S+D)/2

      Where, RPD  =  Relative Percent Difference
               S  =  First Sample Value (original)
               D  =  Second Sample  Value (duplicate)

     The,results  of the duplicate sample analyses must be  reported on FORM
     VI-IN in ug/L for aqueous samples  and mg/Kg dry weight  basis  for solid
     original and duplicate  samples.  A control  limit of 20% for RPD shall
     be used for  original and duplicate sample values greater than or equal
     to 5x CRDL (Exhibit C).  A control  limit  of  (+)  the  CRDL must  be used
     for sample values less  than 5x CRDL,  and the absolute value of the
     control limit (CRDL) must be entered  in  the "Control  Limit" column on
     FORM VI-IN.

     If one result is above  the  5x  CRDL level and the other  is below, use
     the + CRDL criteria.  If both  sample  values are less  than the IDL,  the
     RPD is not calculated on FORM  VI-IN.   For solid sample  or duplicate
     results < 5x CRDL, enter the absolute value of the  CRDL,  corrected for
     sample weight and percent solids,  in  the "Control Limit"  column.
o
 EPA may require additional duplicate sample analyses,  upon Administrative
 Project Officer request, for which the Contractor will be paid.

                                   E-18                                ILM01.0

-------
     If the duplicate sample results are outside the control limits, flag
     all the data for samples received associated with that duplicate sample
     with an "*" on FORMs I-IN and VI-IN.  In the instance where there is
     more than one duplicate sample per SDG,  if one duplicate result is not
     within contract criteria, flag all samples of the same matrix,
     concentration, and method in the SDG.   The percent difference data will
     be used by EPA to evaluate the long-term precision of the methods for
     each parameter.  Specific control limits for each element will be added
     to FORM VI-IN at a later date based on these precision results.

8.   Laboratory Control Sample (LCS) Analysis

     Aqueous and solid Laboratory Control Samples (LCS) must be analyzed for
     each analyte using the same sample preparations,  analytical methods and
     QA/QC procedures employed for the EPA samples received.  The aqueous
     LCS solution must be obtained from EPA (if unavailable, the Initial
     Calibration Verification Solutions may be used).   One aqueous LCS must
     be prepared and analyzed for every group of aqueous samples in a Sample
     Delivery Group, or for each batch of aqueous samples digested,
     whichever is more frequent.   An aqueous  LCS is not required for mercury
     and cyanide analysis.

     The EPA-provided solid LCS must be prepared and analyzed using each of
     the procedures applied to the solid samples received (exception:
     percent solids determination not required).   If the EPA solid LCS is
     unavailable, other EPA Quality Assurance Check samples or other
     certified materials may be used.   One  solid LCS  must be prepared and
     analyzed for every group of solid samples in a Sample Delivery Group,
     or for each batch of samples digested, whichever  is more frequent.

     All LCS results and percent recovery (%R)  will be  reported on FORM VII-
     IN.  If the percent recovery for the aqueous LCS  falls outside the
     control limits of 80-120% (exception:  Ag and Sb),  the analyses must be
     terminated,  the problem corrected,  and the  samples associated with  that
     LCS redigested and reanalyzed.

     If the results for the solid LCS fall  outside the  control limits
     established by EPA,  the analyses must  be terminated,  the problem
     corrected,  and the samples associated  with  that  LCS redigested and
     reanalyzed.

9.   ICP Serial  Dilution Analysis CD

     Prior  to  reporting concentration data  for  the analyte elements,  the
     Contractor  must analyze and  report  the results of  the ICP Serial
     Dilution  Analysis.   The ICP  Serial  Dilution  Analysis  must be performed
     on a sample  from each  group  of  samples of a  similar matrix type (i.e.,
     water,  soil)  and concentration  (i.e.,  low, medium)  or for each Sample
     Delivery  Group,  whichever is more frequent.   Samples  identified as
     field  blanks  cannot be used  for Serial Dilution Analysis.

     If the  analyte concentration is sufficiently high  (minimally a factor
     of 50  above  the instrumental detection limit in the  original sample),
     the serial  dilution (a five  fold dilution) must then  agree  within 10%
     of the  original determination after correction for  dilution.   If the

                                   E-19                               ILM01 0

-------
     dilution analysis  for one  or more  analytes  is not  at  or within 10%,  a
     chemical or physical  interference  effect  must be suspected,  and the
     data for all affected analytes in  the  samples received associated with
     that serial dilution  must  be flagged with an  "E" on FORM  IX-IN and FORM
     I-IN.

     The percent differences for each component  are  calculated as follows:

                             II  -  SI
            % Difference -    	L  x 100
                                 I

     where,   I - Initial Sample Result
             S - Serial Dilution Result  (Instrument Reading x 5)

     In the  instance where there is more  than  one  serial dilution per SDG,
     if one  serial dilution result is not within contract  criteria,  flag  all
     the samples of the same matrix and concentration in the Sample Delivery
     Group.   Serial dilution results and  "E" flags must be reported on FORM
     IX-IN.

10.   Instrument Detection  Limit (IDL) Determination

     Before  any field samples are analyzed  under this contract, the
     instrument detection  limits (in ug/L)  must  be determined  for each
     instrument used, within 30 days of the start  of contract  analyses and
     at least quarterly (every  3 calendar months), and  must meet  the levels
     specified in Exhibit  C.

     The Instrument Detection Limits (in  ug/L)  shall be determined by
     multiplying by 3,  the average of the standard deviations  obtained on
     three nonconsecutive  days  from the analysis of  a standard solution
     (each analyte in reagent water) at a concentration 3x-5x  the instrument
     manufacturer's suggested IDL, with seven  consecutive  measurements per
     day. Each measurement must be performed  as though it were a separate
     analytical sample  (i.e., each measurement must  be  followed by a rinse
     and/or  any other procedure normally  performed between the analysis of
     separate samples).  IDL's  must be  determined  and reported for each
     wavelength used in the analysis of the samples.

     The quarterly determined IDL for an  instrument  must always be used as
     the IDL for that instrument during that quarter.   If  the  instrument  is
     adjusted in anyway that may affect the IDL, the IDL for  that instrument
     must be redetermined and the results submitted  for use as the
     established IDL for that instrument  for the remainder of  the quarter.

     IDLs must be reported for  each instrument used  on  FORM X-IN  submitted
     with each data package.  If multiple AA instruments are used for the
     analysis of an element within a Sample Delivery Group,  the highest IDL
     for the AAs must be used for reporting concentration  values  for that
     Sample  Delivery Group.  The same reporting  procedure  must be used for
     multiple ICPs.
                                   E-20                       '         ILM01.0

-------
11.   Interelement Corrections for ICP

     Before any field samples are analyzed under this contract,  the ICP
     interelement correction factors must be determined prior to the start
     of contract analyses and at least annually thereafter.   Correction
     factors for spectral interference due to Al,  Ca, Fe,  and Mg must be
     determined for all ICP instruments at all wavelengths used  for each
     analyte reported by ICP.  Correction factors  for spectral interference
     due to analytes other than Al, Ca, Fe, and Mg must be reported if they
     were applied.

     If the instrument was adjusted in anyway that may affect the ICP
     interelement correction factors, the factors  must be  redetermined and
     the results submitted for use.  Results from  interelement correction
     factors determination must be reported on FORM XI(PART 1)-IN and FORM
     XI(PART 2)-IN for all ICP parameters.

12.   Linear Range Analysis (LRA)

     For all ICP analyses, a linear range verification check standard must
     be analyzed and reported quarterly (every 3 calendar  months) for each
     element on FORM XII-IN.   The standard must be analyzed during a routine
     analytical run performed under this contract.  The analytically
     determined concentration of this standard must be within +  5% of the
     true value.  This concentration is the upper  limit of the ICP linear
     range beyond which results cannot be reported under this contract
     without dilution of the analytical sample.

13.   Furnace Atomic Absorption (AA) QC Analyses

     Because of the nature of the Furnace AA technique,  the  special
     procedures summarized in Figure 1-Furnace AA  Analysis Scheme ("MSA
     Tree") will be required for quantitation.  (These procedures do not
     replace those in Exhibit D of this SOW, but supplement the  guidance
     provided therein.)

          a. All furnace analyses must fall within the calibration range.
          In addition,  all analyses,  except during full  methods  of Standard
          Addition (MSA),  will require duplicate injections.   The absorbance
          or concentration of each injection must  be reported in the raw
          data as  well  as the average absorbance or concentration values and
          the relative  standard deviation (RSD)  or coefficient of variation
          (CV).   Average concentration values are  used for reporting
          purposes.   The Contractor must be consistent per method and SDG in
          choosing absorbance or concentration to  evaluate which route is to
          be followed in the  MSA Tree.   The Contractor must  also indicate
          which of the  two is being used if both absorbance  and
          concentration are reported in the raw data.  For MSA analysis,  the
          absorbance of each injection must be included  in the raw data.   A
          maximum of 10 full  sample analyses to a  maximum  20 injections may
          be performed between each consecutive calibration  verifications
          and blanks.   For concentrations greater  than CRDL,  the duplicate
          injection readings  must agree within 20% RSD or  CV,  or the
          analytical sample must be rerun once (i.e., two  additional burns).
          If the readings are still out, flag the  value  reported on FORM I-

                                   E-21                                ILM01.0

-------
          IN with an "M".   The  "M"  flag  is  required  for  the  analytical spike
          as well as the  sample.   If  the analytical  spike  for  a  sample
          requires an "M"  flag,  the flag must  be  reported  on FORM I-  IN for
          that sample.

          All furnace analyses  for each  analytical sample, including  those
          requiring an "M" flag,  will require  at  least an  analytical  spike
          to determine if  the MSA will be required for quantitacion.   The
          analytical spike  will  be required to be at a  concentration (in
          the sample) 2x  CRDL  (except for lead which must  be at  20 ug/L).
          This requirement for  an analytical spike will  include  the LCS and
          the preparation  blank.   (The LCS  must be quantitated from the
          calibration curve and corrective  action, if needed,  taken
          accordingly.  MSA is  not to be performed on the  LCS  or preparation
          blank,  regardless of  spike  recovery  results.)  If  the  preparation
          blank analytical spike  recovery is out  of  control  (85-115%) ,  the
          spiking solution must be  verified by respiking and rerunning the
          preparation blank once.   If the preparation blank  analytical spike
          recovery is still out of control,  correct  the  problem  and
          reanalyze all analytical  samples  associated with that  blank.   An
          analytical spifce is not required  on  the pre-digestion  spike
          sample.
          The analytical spike  of a sample  must be run immediately after
          that sample.  The percent recovery (%R) of the spike,  calculated
          by the  same formula as  Spike Sample  Analyses (see  item 6, this
          section),  will then determine  how the sample will  be quantitated,
          as follows:
          1)   If  the  spike  recovery  is  less than 40%,  the sample must  be
               diluted and rerun  with  another  spike.  Dilute the  sample by a
               factor of 5  to 10  and  rerun.  This step must  only be
               performed once.  If  after  the dilution the  spike  recovery is
               still <40% , report data and  flag with an  "E"  to indicate
               interference  problems.

          2)   If  the spike  recovery  is  greater than or  equal  to 40% and the
               sample absorbance  or concentration is less  than 50% of  the
               "spike" , report the sample  results to the  IDL.    If the  spike
               recovery is less than 85%  or  greater  than 115%,  flag the
               result with a "W".
          3)   If  the sample absorbance or concentration is  greater than or
               equal  to 50%  of the  spike  and the spike recovery  is at  or
               between 85% and 115%, the  sample must be quantitated directly
               from the calibration curve and  reported down  to the IDL.
 Analytical Spikes are post-digestion spikes to be prepared prior to
 analysis by adding a known quantity of the analyte to an aliquot of the
 digested sample.  The unspiked sample aliquot must be compensated for any
 volume change in the spike samples by addition of deionized water to the
 unspiked sample aliquot.  The volume of the spiking solution added must
 not exceed 10% of the analytical sample volume; this requirement also
 applies to MSA spikes.

"'"Spike" is defined as [absorbance or concentration of spike sample] minus
 [absorbance or concentration of the sample].

                                   E-22                                ILM01.0

-------
     4)    If  the  sample  absorbance  or  concentration is  greater  than or
          equal to  50% of  the  spike and  the spike  recovery  is less  than
          85%  or  greater than  115%,  the  sample must be  quantitated  by
          MSA.

c.   The  following  procedures  will  be  incorporated into MSA analyses.

     1)    Data from MSA  calculations must be within the  linear  range  as
          determined by  the calibration  curve generated  at  the
          beginning of the analytical  run.

     2)    The  sample and three spikes  must be analyzed consecutively
          for MSA quantitation (the  "initial" spike run  data is
          specifically excluded from use in the MSA quantitation).
          Only single injections are required for MSA quantitation.

          Each full MSA  counts as two  analytical samples towards
          determining 10% QC frequency (i.e., five  full  MSAs can be
          performed between calibration verifications).

     3)    For analytical runs containing only MSAs,  single  injections
          can be  used for QC samples during that run.  For  instruments
          that operate in an MSA mode only, MSA can be used to
          determine QC samples during  that run.

     4)    Spikes  must be prepared such that:
          a)  Spike  1  is  approximately  50%  of  the  sample concentration.
         b)  Spike  2  is  approximately  100% of  the  sample
             concentration.

          c)  Spike  3  is  approximately  150% of  the  sample
             concentration.

     5)   The data for each MSA analysis must be clearly identified in
          the raw data documentation (using added concentration as the
         x-variable and absorbance as the y-variable) along with the
         slope,  x-intercept,  y-intercept and correlation coefficient
          (r)  for the least squares fit of the data.  The results must
         be reported on FORM VIII-IN.   Reported values obtained by MSA
         must be flagged on the  data sheet (FORM I-IN)  with the letter
          "S"  if the correlation  coefficient  is greater than or  equal
         to 0.995.

    6)   If the  correlation coefficient (r)  for a particular analysis
         is less than 0.995,  the  MSA analysis  must be repeated  once.
         If the  correlation coefficient is still  less than 0.995,
         report  the results on FORM I-IN from the run with the  best
         "r"  and flag the  result  with  a "+"  on FORM VIII-IN and FORM
         I-IN.
                             E-23                                ILM01 0

-------
                                        Figure 1.
                          FURNACE ATOMIC ABSORPTION ANALYSIS  SCHEME
      PREPARE AND ANALYZE
     SAMPLE AND ONE SPIKE
          (2  X CRDL)
   (Double  Injections  Required)
       ANALYSES WITHIN
      CALIBRATION RANGE
                YES
      RECOVERY OF SPIKE
        LESS  THAN 40%
               NO
    SAMPLE  ABSORBANCE OR
   CONCENTRATION LESS^ THAN
   50% OF SPIKE ABSORBANCE
       OR CONCENTRATION
               NO
        SPIKE  RECOVERY
       LESS THAN  85%  OR
       GREATER THAN 115%
                 YES
  QUANTITATE BY MSA WITH 3
  SPIKES  AT 50, 100 £ 1SO\
    OF  SAMPLE ABSORBANCE
      OR  CONCENTRATION
(Only Single  Injeclioni Rr
  CORRELATION COEFFICIENT
       LESS THAN 0.995
                 NO
     FLAG  DATA WITH "S"
                                             NO
                                   If YES, Repeat  Only ONCE
                                   If Still YES
                                                     NO
YES
  SPIKE RECOVERY
 LESS  THAN 85% OR
GREATER THAN  115%
                                                     YES
             NO
                                    DILUTE  SAMPLE
                                      AND SPIKE
                                 If YES, Repeat Only ONCE
                                         If Still VES
                                                                    FLAG DATA
                                                                   WITH AN "E"
                                                                 REPORT RESULTS
                                                                  DOWN TO  IDL
                                   REPORT  RESULTS
                                    DOWN TO  IDL,
                                   FLAG WITH A "W"
                            QUANTITATE  FROM
                           CALIBRATION  CURVE
                            AND REPORT  DOWN
                                 TO IDL
                                      FLAG DATA
                                     WITH A  -*••
                                      E-24
                                                                            ILM01.0

-------
                                    SECTION VI

                           LABORATORY  EVALUATION PROCESS

      This document outlines the procedures which will be used by the
Administrative Project Officer or his/her authorized representative to
conduct laboratory audits to determine the Contractor's ability to meet the
terms and conditions of this contract.  The evaluation process incorporates
two major steps:  1) evaluation of laboratory performance, and 2) on-site
inspection of the laboratory to verify continuity of personnel,
instrumentation and quality control requirements of the contract.

I.   Evaluation of Laboratory Performance

     a.    Performance  Evaluation Sample Analysis

          1)   The Performance  Evaluation (PE) sample set  will  be  sent  to a
              participating  laboratory on a quarterly basis to verify  the
              laboratory's continuing ability to produce  acceptable
              analytical results.  These samples will be  provided either
              single blind  (recognizable as a PE material and  of  unknown
              composition),  or  double blind (not recognizable  as  PE
              material and of  unknown composition).  If received  as a
              single blind,  the  Contractor is required to submit  PE sample
              data  in a separate SDG  package in accordance with Delivery
              Schedule requirements for  PE Sample  data.   PE samples
              received as double blind would be treated as routine samples
              and data would be  submitted in the SDG deliverables package
              per normal procedure.

          2)   When  the PE  data are received by EPA, results will  be scored
              routinely for  identification and quantitation.   Results  of
              these scorings will be  provided for  the Contractor  via coded
              evaluation spreadsheets by analyte.  The Government may
              adjust  the scores  on any given PE sample to compensate for
              unanticipated  difficulties with a particular sample.

          3)   If the Contractor  laboratory performs unacceptably, the
              Contractor will  be notified by the Administrative Project
              Officer.  A  laboratory  so  notified may expect, but  the
              Government is  not  limited  to, the following actions:  a  site
              visit, a full  data audit,  cessation  of sample shipments,
              and/or laboratory  analysis of a second PE sample.   Failure by
              the laboratory to  take  corrective actions and/or failure of
              two successive PE  sample analyses is indicative  of  Contractor
              failure to maintain technical competence and will require
              that  the laboratory discontinue analysis of samples until
              such  time as the Administrative Project Officer  has
              determined that  the laboratory has corrected the problem and
              may resume analyses.
     b.    Inorganic  Data Audit

          Inorganic  data audits are conducted by EMSL-LV on the Contractor's
          sample  data  packages.   The inorganic data audit  provides the
          Agency  with  an  in-depth inspection and evaluation of  the data
          packages with regard  to achieving QA/QC acceptability.

                                   E-25                                ILM01 0

-------
2.    On-Sice Laboratory Evaluation

      The on-site laboratory evaluation helps to ensure that technical
      competence is maintained and that all the necessary quality control
      is being applied by the Contractor in order to deliver a quality
      product.

     a.   Or.-site  laboratory  evaluations allow  the evaluators  to  determine
         that:
         1)   The organization and personnel are qualified to perform
              assigned  tasks;

         2)   Adequate  facilities and equipment are available;

         3)   Complete  documentation, including chain-of-custody of samples
              is being  implemented;

         4)   Proper analytical methodology is being used;

         5)   Adequate  analytical quality control, including reference
              samples,  control charts, and documented corrective action
              measure:., is being provided; and

         6)   Acceptable data handling and documentation techniques are
              being used.

     b.   The  on-site visit also serves as a mechanism for discussing
         weaknesses identified through Performance Evaluation sample
         analysis or through Contract Compliance Screening or other review
         of  data deliverables.  Lastly, the on-site visit allows the
         evaluation team to  determine if the laboratory has implemented  the
         recommended and/or  required corrective actions, with respect to
         quality assurance,  that were made during the previous on-site
         visit.
                                   E-26                                ILM01.0

-------
             EXHIBIT F
CHAIN-OF-CUSTODY, DOCUMENT CONTROL,
 AND STANDARD OPERATING PROCEDURES
              F-l                               ILM01.0

-------
1.    SAMPLE CHAIN-OF-CUSTODY

      A sample is physical evidence collected from a facility or from the
      environment.  Controlling evidence is an essential part of the
      hazardous waste investigation effort.  To accomplish this, Contractors
      are required to develop and implement the following sample
      identification, chain-of-custody, sample receiving, and sample tracking
      procedures.

1.1   Sample Identification

      To assure traceability of the samples while in possession of the
      Contractor, the Contractor shall have a specified method for
      maintaining identification of samples throughout the laborcitory.

      Each sample and sample preparation container shall be labeled with the
      EPA number or a unique laboratory identifier.  If a unique laboratory
      identifier is used, it shall be cross-referenced to the EPA number.

1.2   Chain-of-Custody Procedures

      Because of the nature of the data being collected, the custody of EPA
      samples must be traceable from the time the samples are collected until
      they are introduced as evidence in legal proceedings.  The Contractor
      shall have procedures ensuring that EPA sample custody is maintained
      and documented.  A sample is under custody if:

      o  It  is  in your possession,  or

      o  It  is  in your view  after  being in  your possession, or
      o  It  was  in your  possession and  you  locked  it up,  or
      o  It  is  in a designated  secure area.   (Secure areas  shall be  accessible
         only to  authorized  personnel.)

1.3   Sample Receiving Procedures

      1.3.1    The Contractor shall designate  a sample custodian responsible
               for receiving all samples.

      1.3.2    The Contractor shall designate  a representative  to  receive
               samples in the event that the  sample custodian is not
               available.

      1.3.3    The condition of the shipping  containers  and sample bottles
               shall be  inspected  upon  receipt by  the sample  custodian or
               his/her representative.
                *
      1.3.4    The condition of the custody seals  (intact/not intact)  shall  be
               inspected upon receipt by the  sample custodian or his/her
               representative.

      1.3.5    The sample custodian or  his/her representative shall  check for
               the presence  or  absence  of the following  documents  accompanying
               the sample shipment:

                                    F-2                               ILM01.0

-------
               o  Airbills or airbill stickers
               o  Custody seals
               o  EPA custody records
               o  EPA traffic reports or SAS packing lists
               o  Sample tags
      1.3.6    The  sample custodian  or his/her representative  shall  sign and
               date all  forms  (e.g.,  custody  records,  traffic  reports  or
               packing  lists,  and airbills) accompanying  the samples at  the
               time of  sample  receipt.
      1.3.7    The  Contractor  shall  contact the Sample Management Office (SMO)
               to resolve discrepancies and problems such as absent  documents,
               conflicting  information, broken custody seals,  and
               unsatisfactory  sample condition (e.g., leaking  sample bottle).
      1.3.8    The  Contractor  shall  record the resolution of discrepancies  and
               problems  on  Telephone Contact  Logs.
      1.3.9    The  following information shall be recorded on  Form DC-1  (See
               Exhibit B) by the sample custodian or his/her representative as
               samples are  received  and inspected:
               o  Condition of the  shipping container
               o 'Presence or absence and condition  of custody seals  on
                 shipping and/or sample containers
               o  Custody seal numbers,  when present
               o  Condition of the  sample  bottles
               o  Presence or absence of airbills  or airbill  stickers
               o  Airbill or airbill  sticker numbers
               o  Presence or absence of EPA custody records
               o  Presence or absence of EPA traffic reports  or SAS packing
                 lists
               o  Presence or absence of sample tags
               o  Sample tag identification numbers  cross-referenced  to  the
                 EPA sample numbers
               o  Verification of agreement or non-agreement  of information
                 recorded on shipping documents and sample containers
               o  Problems or discrepancies
               i
1.4   Sample Tracking Procedures
      The Contractor shall maintain records documenting all  phases  of sample
      handling from receipt to final analysis.
                                   F-3                                ILM01.0

-------
2.    DOCUMENT CONTROL PROCEDURES

      The goal of the laboratory document control program is to assure that
      all documents for a specified Sample Delivery Group (SDG) will be
      accounted for when the project is completed.  Accountable documents
      used by contract laboratories shall include, but not be limited to,
      logbooks, chain-of-custody records, sample work sheets, bench sheets,
      and other documents relating to the sample or sample analyses.  The
      following document control procedures have been established to assure
      that all laboratory records are assembled and stored for delivery to
      EPA or are available upon request from EPA prior to the delivery
      schedule.

2.1   Preprinted Laboratory Forms and Logbooks

      2.1.1   All documents produced by  the Contractor which are directly
              related to the preparation and analysis of EPA samples shall
              become the property of the EPA and shall be placed in the
              complete sample delivery group file  (CSF)   All observations
              sr.d results recorded by the laboratory but not on preprinted
              laboratory forms shall be  entered into permanent laboratory
              logbooks.  When all data from a SDG  is compiled, all original
              laboratory forms and copies of all SDG-related logbook entries
              shall be included in the documentation package.

      2.1.2   The Contractor shall identify the activity recorded on all
              laboratory documents which are directly related to the
              preparation and analysis of EPA samples.

      2.1.3   Pre-printed laboratory forms shall contain the name of the
              laboratory and be dated (month/day/year) and signed by the
              person responsible for performing the activity at the time an
              activity is performed.

      2.1.4   Logbook entries shall be dated (month/day/year) and signed by
              the person responsible for performing the activity at the time
              an activity is performed.

      2.1.5   Logbook entries shall be in chronological order.  Entries in
              logbooks, with the exception of instrument run logs and
              extraction logs, shall include only one SDG per page.

      2.1.6   Pages in both bound and unbound logbooks shall be sequentially
              numbered.

      2.1.7   Instrument run logs shall be maintained so as to enable a
              reconstruction of the run sequence of individual instruments.

              Because the laboratory must provide copies of the instrument
              run logs to EPA, the laboratory may exercise the option of
              using only laboratory or EPA sample identification numbers in
              the logs for sample ID rather than government agency or
              commercial client names to preserve the confidentiality of
              commercial clients.

                                   F-4                               ILM01.0

-------
      2.1.8    Corrections to supporting documents and raw data shall be made
               by drawing a single line through the error and entering the
               correct information.  Corrections and additions to supporting
               documents and raw data shall be dated and initialed.   No
               information shall be  obliterated or rendered unreadable.

               All notations shall be recorded in ink.

               Unused portions of documents shall be "z'd"  out.

2.2   Consistency of Documentation

      The Contractor shall assign a  document control officer responsible  for
      the organization and assembly  of the CSF.

      All copies of laboratory documents shall be complete and legible.

      Original documents which include information relating to more than  one
      SDG shall be filed in the CSF  of the lowest SDG number.  The copy(s)
      shall be placed in the other CSF(s) and the Contractor shall record the
      following information on the copy(s) in red ink:

      "COPY

      ORIGINAL IS FILED IN CSF 	"
      The Contractor shall sign and date this addition to the copy(s).

      Before releasing analytical results, the document control officer shall
      assemble and cross-check the information on samples tags, custody
      records, lab bench sheets, pers. -.al ar.l instrument logs,  and other
      relevant deliverables to ensure :hat data pertaining to each particular
      sample or sample delivery group Is consistent throughout  the CSF,

2.3   Document Numbering and Inventory Procedure

      In order to provide document accountaoility of the completed analysis
      records*, each item in the CSF shall be inventoried and assigned a
      serialized number as described in Exhibit B).

      All documents relevant to each sample delivery group,  including logbook
      pages,  bench sheets, mass spectra, chromatograms,  screening records,
      re-preparation records,  re-analysis records, records of failed or
      attempted analysis,  custody records, library research results,  etc.
      shall  be inventoried.

      The Document Control Officer (DCO) shall be responsible for ensuring
      that all documents generated are placed in the CSF for inventory and
      are delivered to the appropriate EPA region or other receiver as
      designated by EPA.  The  DCO shall place the sample tags in plastic bags
      in the  file.
                                   F-5                               ILM01.0

-------
2.4   Storage of EPA Files

      The Contractor shall maintain EPA laboratory documents in a secure
      location.

2.5   Shipment of Deliverables

      The Contractor shall document shipment of deliverables packages to the
      recipients.  These shipments require custody seals on the containers
      placed such that they cannot be opened without damaging or breaking the
      seal.  The Contractor shall document what was sent,  to whom, the date,
      and the method (carrier) used.

      A copy of the transmittal letter for the CSF shall be sent to the
      NEIC/CEAT and the SMO.

3.     SPECIFICATIONS FOR WRITTEN STANDARD OPERATING PROCEDURES

      The Contractor shall have written standard operating procedures (SOPs)
      for receipt of samples,  maintenance of custody,  sample identification,
      sample storage, sample tracking, and assembly of completed data.

      An SOP is defined as a written narrative stepwise description of
      laboratory operating procedures including examples of laboratory
      documents.  The SOPs shall accurately describe the actual procedures
      used in the laboratory,  and copies of the written SOPs shall be
      available to-the appropriate laboratory personnel.   These procedures
      are necessary to ensure that analytical data produced under this
      contract are acceptable for use in. EPA enforcement case preparation and
      litigation.  The Contractor's SOPs shall provide mechanisms and
      documentation to meet each of the following specifications and shall be
      used by EPA as the basis for laboratory evidence audits.

3.1   The Contractor shall have written SOPs describing the sample
      custodian's duties and responsibilities.

3.2   The Contractor shall have written SOPs for receiving and logging in of
      the samples.  The procedures shall include but not be limited to
      documenting the following information:

      3.2.1    Presence or  absence  of  EPA chain-of-custody  forms

      3.2.2    Presence or  absence  of  airbills or airbill stickers

      3.2.3    Presence or  absence  of  traffic  reports or SAS packing  lists

      3.2.4    Presence or  absence  of  custody  seals  on  shipping and/or sample
               containers and  their condition

      3.2.5    Custody seal numbers, when present

      3.2.6    Airbill  or airbill  sticker numbers

      3.2.7    Presence or  absence  of  sample tags

                                    F-6                               ILM01.0

-------
      3.2.8    Sample  tag ID numbers

      3.2.9    Condition  of the  shipping container

      3.2.10   Condition  of the  sample bottles

      3.2.11   Verification of agreement or non-agreement of  information on
               receiving  documents and sample containers

      3.2.12   Resolution of problems or discrepancies with the SMO

      3.2.13   An explanation of any terms used by  the laboratory  to describe
               sample  condition  upon receipt (e.g., good, fine, OK)

3.3   The Contractor shall have written SOPs for maintaining identification
      of EPA samples throughout the laboratory.

      If the Contractor assigns unique laboratory identifiers,  written SOPs
      shall include a description of the method used to assign the unique
      laboratory identifier and shall include  a description of the document
      used to cross-reference the unique laboratory identifier to the EPA
      sample number.

      If the Contractor uses  prefixes or suffixes in addition to sample
      identification numbers,  the written SOPs shall include their
      definitions.

3,4   The Contractor shall have written SOPs describing all storage areas for
      samples in the laboratory.  The SOPs shall include a. list of authorized
      personnel who have access or keys to secure storage areas.

3.5   The Contractor shall have written SOPs describing the method by which
      the laboratory maintains samples under custody.

3.6   The Contractor shall have written SOPs describing the method by which
      the laboratory maintains the security of any areas identified as
      secure.

3.7   The Contractor shall have written SOPs for tracking the work performed
      on any particular samples.  The tracking SOP shall include:

      o A description of the documents used to record sample receipt, sample
        storage, sample  transfers, sample preparations, and sample analyses.

      o A description of the documents used to record calibration and QA/QC
        laboratory work.
      o Examples of document formats and laboratory documents used in the
        sample receipt,  sample  storage, sample transfer, and sample analyses.

      o A narrative step-wise description of how documents are used to track
        samples.

3.8   The Contractor shall have written SOPs for organization and assembly of
      all documents relating to each SDG.   Documents  shall  be filed on a
      sample delivery group-specific basis.  The procedures shall ensure that

                                   F-7                                ILM01.0

-------
      all documents including logbook pages, sample tracking records,
      chromatographic charts, computer printouts, raw data summaries,
      correspondence, and any other written documents having reference to the
      SDG are compiled in one location for submission to EPA.   The written
      SOPs shall include:

      o A description of the numbering and  inventory method.
      o A description of the method used by the laboratory to verify
        consistency  and  completeness of the CSF.

      o Procedures for the shipment of deliverables packages using custody
        seals.
4.    HANDLING OF CONFIDENTIAL INFORMATION

      A Contractor conducting work under this  contract may receive EPA-
      designated confidential information from the agency.   Confidential
      information must be handled separately from other documentation
      developed under this contract.   To accomplish this,  the  following
      procedures for  the  handling of confidential information  have been
      established.

4.1   All confidential documents shall be  under the supervision of a
      designated Document Control Officer  (DCO).

4.2   Confidential Information

      Any samples or  information received  with a  request of confidentiality
      shall  be handled as "confidential."  A separate  locked file shall be
      maintained to store this information and shall be segregated from other
      nonconfidential information.   Data generated from confidential samples
      shall  be treated as confidential.  Upon  receipt  of confidential
      information,  the DCO will log these  documents into a  Confidential
      Inventory Log.   The information will then be available to  authorized
      personnel but only  after it has been signed out  to that  person by the
      DCO.  The documents shall be returned  to the locked file at the
      conclusion of each  working day.   Confidential information may not be
      reproduced except upon approval by the EPA  Administrative  or Technical
      Project Officer.  The DCO will  enter all copies  into  the document
      control system  described above.   In  addition,  this information may not
      be disposed of  except upon approval  by the  EPA Administrative or
      Technical Project Officer.   The DCO  shall remove  and  retain the cover
      page of any confidential  information disposed of  for  one year and shall
      keep a record on the disposition in  the  Confidential  Inventory Log.
                                   F-8                               ILM01.0

-------
    EXHIBIT G
GLOSSARY OF TERMS
                                      ILM01.0

-------
                              GLOSSARY OF TERMS
 ABSORBANCE -  a measure of the  decrease in incident  light passing  through a
 sample  into the detector.   It  is  defined mathematically as:

               ,  _  I ("solvent)   _  log lo
                     I(solution)          I

                Where,  I - radiation  intensity

 ALIQUOT -  a measured portion of a field sample  taken  for analysis.

 ANALYSIS DATE/TIME  - the date  and military time  (24-hour clock) of  the
 introduction  of the sample, standard,  or blank  into the analysis  system.

 ANALYTE -  the element or ion an analysis seeks  to determine; the  element of
 interest.

 ANALYTICAL SAMPLE -  Ar_/ solution  or media introduced  ir.to an instrument  on
 which an analysis is performed excluding instrument calibration,  initial
 calibration verification,  initial  calibration blank,  continuing calibration
 verification  and continuing calibration blank.  Note  the following  are all
 defined as  analytical  samples:  undiluted and diluted samples (EPA  and non-
 EPA), predigestion  spike samples,  duplicate samples,  serial dilution
 samples, analytical  spike  samples, post-digestion spike samples,
 interference  check  samples  (IGS),  CRDL standard for AA (CRA), CRDL  standard
 for  ICP (CRI),  laboratory  control  sample (LCS), preparation blank (PB) and
 linear  range  analysis  sample (LRS).

 ANALYTICAL SPIKE -  The  furnace post-digestion spike.  The addition  of a
 known amount  of  standard after digestion.

 AUTOZERO -  zeroing  the  instrument  at the proper wavelength.  It is
 equivalent  to  running  a  standard blank with the absorbance set at zero.

AVERAGE INTENSITY -  The  average of two different injections (exposures).

BACKGROUND  CORRECTION  -  a technique to compensate for variable background
contribution to  the  instrument signal in the determination of trace
elements.

BATCH - a group  of samples prepared at the same time  in the same location
using the same method.

CALIBRATION - the establishment of an analytical curve based on the
absorbance, emission intensity, or other measured characteristic of known
standards.   The  calibration standards must be prepared using the same type
of acid or concentration of acids  as used in the sample preparation.

CALIBRATION BLANK -  a volume of acidified deionized/distilled water.

CASE -  a finite, usually predetermined number of samples  collected over a
given time period from a particular site.  Case  numbers are assigned by the


                                   G-l                                 ILM01.0

-------
 Sample Management Office.   A Case  consists of one  or more  Sample  Delivery
 Groups.

 COEFFICIENT OF VARIATION (CV)  -  the  standard deviation as  a percent of the
 arithmetic mean.

 CONCENTRATION LEVEL (low or medium)  -  for  inorganics analysis,  low or
 medium level is defined  by  the appropriate designation checked  by the
 sampler  on the Traffic Report.

 CONTINUING CALIBRATION -  analytical  standard run every 10  analytical
 samples  or every 2  hours, whichever  is more frequent,  to verify the
 calibration of the  analytical  system.

 CONTRACT REQUIRED DETECTION LIMIT  (CRDL) -  minimum level of detection
 acceptable under the contract  Statement of Work.

 CONTROL  LIMITS  -  a  range within which  specified measurement results must
 fall  to  be compliant.  Control limits  may  be  mandatory, requiring
 corrective action if exceeded, or  advisory,  requiring  that  noncompliant
 data  be  flagged.

 CORRELATION COEFFICIENT  - a number (r) which  indicates  the  degree of depen-
 dence between two variables (concentration  -  absorbance).   The  more
 dependent  they are  the closer  the value to  one.  Determined on  the basis  of
 the least  squares line.

 DAY -  unless otherwise specified,  day  shall mean calendar day.

 DIGESTION  LOG -  an  official record of  the  sample preparation  (digestion).

 DISSOLVED  METALS  -  analyte  elements which have not been digested  prior to
 analysis and which  will pass through a 0.45 um filter.

 DRY WEIGHT -  the  weight of  a sample based on percent solids.  The weight
 after  drying in an  oven.

 DUPLICATE  -  a second aliquot of a sample that is treated the same  as  the
 original sample  in  order to determine  the precision  of the method.

 FIELD  BLANK  - any sample  submitted from the field identified as a blank.

 FIELD  SAMPLE  - a  portion of material received to be analyzed that  is
 contained  in  single or multiple containers and identified by a unique EPA
 Sample Number.

 FLAME ATOMIC ABSORPTION (AA) - atomic absorption which utilizes flame for
 excitation.

GRAPHITE FURNACE ATOMIC ABSORPTION (GFAA) - atomic absorption which
utilizes a graphite cell for excitation.

HOLDING TIME  - the  elapsed  time expressed in days from the date of receipt
of the sample by  the Contractor until  the date of its analysis.

                                   G-2                                ILM01.0

-------
      Holding  time  — (sample  analysis  date  -  sample receipt  date)

 INDEPENDENT  STANDARD -  a  Contractor-prepared  standard solution  that  is
 composed  of  analytes from a different  source  than those  used in the
 standards for  the initial calibration.

 INDUCTIVELY  COUPLED PLASMA (ICP)  - a technique  for the simultaneous  or
 sequential multi-element  determination of elements in solution.  The basis
 of  the method  is the measurement  of atomic  emission by an  optical
 spectroscopic  technique.   Characteristic atomic line emission spectra are
 produced  by  excitation  of the sample in a radio frequency  inductively
 coupled plasma.

 IN-HOUSE  - at  the Contractor's  facility.

 INJECTION -  introduction  of the analytical  sample into the instrument
 excitation system for the purpose of measuring  absorbance, emission  or
 concentration  of an analyte.  May also  be referred to  as exposure.

 INSTRUMENT CALIBRATION  -  analysis of analytical standards  for a  series of
 different specified concentrations; used to define the quantitative
 response, linearity,  and  dynamic  range  of the instrument to  target
 analytes.

 INSTRUMENT DETECTION LIMIT (IDL)  - determined by multiplying by  three the
 standard  deviation  obtained for the analysis of a standard solution  (each
 analyte in reagent  water)  at a  concentration of 3x-5x  IDL on three
 nonconsecutive days  with  seven  consecutive  measurements per  day.

 INTERFERENTS - substances which affect  the  analysis  for  the  element  of
 interest.

 INTERNAL  STANDARDS  -  in-house compounds added at  a known concentration.

 LABORATORY - synonymous with Contractor as  used herein.

 LABORATORY CONTROL  SAMPLE  (LCS) - a control sample of  known  composition.
 Aqueous and solid laboratory control samples are  analyzed using the  same
 sample preparation,  reagents,  and analytical methods employed for the EPA
 samples received.

 LABORATORY RECEIPT  DATE -  the date on which a sample is received at  the
 Contractor's facility, as recorded on the shipper's delivery receipt and
 sample Traffic Report.  Also referred to as VTSR  (validated  time of  sample
 receipt).

 LINEAR RANGE, LINEAR DYNAMIC RANGE - the concentration range over which the
 ICP analytical curve remains linear.

MATRIX -  the predominant material of which  the  sample  to be  analyzed is
 composed.   For the  purpose of this SOW, a sample matrix is either water or
 soil/sediment.  Matrix  is not synonymous with phase (liquid  or solid).
                                   G-3                                 ILM01.0

-------
MATRIX MODIFIER - salts used in AA to lessen the effects of chemical
interferents, viscosity, and surface tension.

MATRIX SPIKE  - aliquot of a sample (water or soil) fortified (spiked) with
known quantities of specific compounds and subjected to the entire
analytical procedure  in order to indicate the appropriateness of the method
for  the matrix by measuring recovery.

METHOD OF STANDARD ADDITIONS (MSA) - the addition of 3 increments of a
standard solution (spikes) to sample aliquots of the same size.
Measurements  are made on the original and after each addition.  The slope,
x-intercept and y-intercept are determined by least-square analysis.  The
analyte concentration is determined by the absolute value of the x-
intercept.  Ideally,  the spike volume is low relative to the sample volume
(approximately 10% of the volume).  Standard addition may counteract matrix
effects; it will not counteract spectral effects.  Also referred to as
Standard Addition.

PERCENT SOLIDS - the proportion of solid in a soil sample determined by
drying an aliquot of the sample.

PERFORMANCE EVALUATION (PE) SAMPLE - a sample of known composition provided
by EPA for Contractor analysis.   Used by EPA to evaluate Contractor
performance.

PREPARATION BLANK (reagent blank, method blank) - an analytical control
that contains distilled, deionized water and reagents,  which is carried
through the entire analytical procedure (digested and analyzed).   An
aqueous method blank is treated with the same reagents  as a sample with a
vater matrix; A solid method blank is treated with the  same reagents as a
soil sample.

PROTOCOL - a  compilation of the procedures to be followed with respect no
sample receipt and handling, analytical methods, data reporting and
deliverables, and document control.  Used synonymously  with Statement of
Work (SOW).
                                                                              t
ROUNDING RULES - If the figure following those to be retained is less than
5, the figure is dropped, and the retained figures are  kept unchanged.  As
an example,  11.443 is rounded off to 11.44.

If the figure following those to be retained is greater than 5, the figure
is dropped,  and the last retained figure is raised by 1.   As an example,
11.446 is rounded off to 11.45.

If the figure following those to be retained is 5,  and  if there are no
figures other than zeros beyond the five,  the figure 5  is dropped,  and the
last-place figure retained is increased by one if it is an odd number or it
is kept unchanged if an even number.   As an example,  11.435 is rounded off
to 11.44,  while 11.425 is rounded off to 11.42.

If a series of multiple operations is to be performed (add,  subtract,
divide,  multiply), all figures are carried through the  calculations.  Then
the final answer is rounded to the proper number of significant figures.


                                   G-4                                 ILM01.0

-------
 See forms instructions (Exhibit B)  for exceptions.

 RUN - a continuous analytical sequence consisting of prepared samples and
 all associated quality assurance measurements  as  required by the  contract
 Statement of Work.

 SAMPLE DELIVERY GROUP (SDG)  - a unit within a  sample Case that is used to
 identify a group of samples  for delivery.   An  SDG is a group of 20 or fewer
 samples within a Case,  received over a period  of  up  to 14 calendar days.
 Data from all samples in an  SDG are  due concurrently.   A Sample Delivery
 Group is defined by one of the following, whichever  occurs first:

        o   Case;  or •
        o   Each  20  samples within a  Case; or
        o   Each  14-day  calendar period during which  samples  in  a Case  are
            received, beginning with  receipt of the first  sample in  the Case
            or  SDG.

 Samples may be assigned to Sample Delivery  Groups by matrix  (i.e.,  all
 soils in one SDG,  all  waters  in another), at the  discretion  of  the
 laboratory.

 SAMPLE NUMBER  (EPA  SAMPLE NUMBER) -  a unique identification  number
 designated by  EPA for  each sample.   The  EPA Sample Number appears on  the
 sample Traffic Report which documents  information on that sample.

 SERIAL DILUTION  -  the dilution  of a  sample by a factor  of five.  When
 corrected by the  dilution factor, the diluted sample must agree with  the
 original undiluted  sample within specified  limits.   Serial dilution may
 reflect the influence of  interferents.

 SOIL -  synonymous with  soil/sediment or  sediment as used  herein.

 STOCK SOLUTION -  a  standard solution which can be diluted to derive other
 standards.

 TOTAL METALS - analyte  elements which have been digested  prior to analysis.

 TRAFFIC  REPORT (TR)  - an  EPA  sample  identification form filled out by  the
 sampler, which accompanies the sample during shipment to  the laboratory and
which  is used  for documenting sample condition and receipt by the
 laboratory.

WET WEIGHT  - the weight of a sample aliquot including moisture (undried).

10% FREQUENCY  -t a frequency specification during an analytical sequence
allowing for no more than 10 analytical samples between required
calibration verification measurements, as specified by the contract
Statement of Work.
                                   G-5                                 ILM01.0

-------

-------
                                  EXHIBIT H

               DATA DICTIONARY AND FORMAT FOR DATA DELIVERABLES
                          IN  COMPUTER-READABLE FORMAT
                                                                      Page No.
SECTION I:      Description of Deliverables 	  H-1
SECTION II:    Format A Specification 	  H-3
SECTION III:   Format B Specification 	  H-26
                                                                  ILM01.0

-------

-------
                                   SECTION I
                          DESCRIPTION OF DELIVERABLES
1.  Introduction

1.1   Two file formats are specified  for delivery of computer-readable data.
      Format A is oriented to  the structure of the hardcopy reporting forms
      required by the contract.  Format B is oriented to the general data
      required by the contract.  Information sufficient to generate required
      hardcopy forms is contained in  either format.  NOTE:  Beginning in
      March, 1991, all data submitted in electronic form shall be submitted
      using the USEPA Standard for Electronic Transmission of Laboratory
      Results, EPA Order 2180.2

1.2   The file or files for a  Sample  Delivery Group (SDG, see Exhibit A,
      Section I, B) must be submitted on a diskette or diskettes (see
      paragraph, 2.1).  Information on a diskette or diskettes for any single
      SDG must be in one, and  only one, of the two formats.  The format used
      13 at the option of the  laboratory.  The option used must be included
      in the File Name specification  (paragraph 2.2).

1.3   Format A consists of variable length ASCII records, and Format B
      consists of fixed-length 80-byte ASCII records.

1.4   All information for one  SDG must be in one file if format A is used.
      Use of Format B may require information for one SDG to be in a number
      of files.   Format B may  require  more than one 360 K diskette for a
      valid SDG.

2.    Deliverable

2.1   The file or files niust be submitted on a 5-1/4 inch floppy diskette,
      which may be either a double-sided, double density,  360 K-byte or a
      high capacity 1.2 M-byte, or 3.5 inch double-sided,  double-density 720
      K-byte or 1.44 M-byte,  diskette.  The diskette must be formatted and
      recorded using the MS-DOS Operating System.   The  diskette or diskettes
  .   must contain all information relevant to one and  only one SDG,  and must
      accompany the hardcopy package  for the  SDG submitted to the Sample
      Management Office (see Exhibit  B).   Information on the  diskette  or
      diskettes  must correspond exactly with  information submitted in  the
      hardcopy data package  and on the hardcopy data package  forms.  Blank  or
      unused records in either format should  not be  included  on the
      diskettes.
                                      H-l                        ILM01.0

-------
2.2   Each diskette must be identified with an external label containing (in
      this order) the following information:

               Disk Density
               File Name(s)
               Laboratory Name (optional)
               Laboratory Code
               Case Number (where applicable)
               SAS Number (where applicable)

      The format for the File Name(s) must be XXXXXX.INY

               where  XXXXXX  is the SDG identifier

                      I       indicates Inorganics analysis

                      N       is a continuation number used to identify
                              multiple files corresponding to the same SDG.
                              For Format A, "N" must be "1".   For Format B,
                              "N" must be "1" for the only  or first file of
                              the SDG, and must be incremented to "2", "3",
                              etc., for subsequent files of the SDG.  "N"
                              cannot be greater than 9.  If "N" is greater
                              than 9 then replace Y with a digit to continue
                              incrementing.  The files must be incremented in
                              chronological order.

                      Y       is "A" for Format A
                              is "B" for Format B, or a digit (0 to 9) if
                              more than 9 Format B files are used.

Examples:

      Format A          ASC123.I1A

      Format 3          ABC123.I1B
                        ABC123.I2B
                        ABC123.I3B
                        ABC123.I9B
                        ABC123.I10
                        ABC123.I11
                        ABC123.I99
      Dimensions of the label must be in the range 4-3/4" to 5" long by 1-1/4
      to 1-1/2" wide.
                                   H-2                                 ILM01.0

-------
                                  SECTION II

                            FORMAT A SPECIFICATION
1.     Format Characteristics

1.1   Format A is based upon the structure of the hardcopy reporting forms
      required by the contract.  With two exceptions, Form Suffix and Record
      Type, all fields in the format correspond directly with entries or
      i terns on the hardcopy forms.

1.2   Format A includes detailed specifications for the required format of
      each Inorganic Reporting Form's HEADER and DETAIL records.  The exact
      columns in which each field is to be contained are shown,  as well as
      the length of the field.   Each field's required contents are specified
      either as a literal (contained in single quotes) which must appear
      exactly as shown (without the quotes),  or as a variable for which a
      format is listed in the format column.   Each field's required format is
      sp.cified either as an option of two or more choices (divided by
      slashes), as MM/DD/YY for a date, as a CHARACTER field,  or as a NUMERIC
      field.

1.3   Format fields listed as CHARACTER may contain any standard ASCII
      characters,  and must be left-justified and padded with blanks.  Formats
      listed as NUMERIC may contain numeric digits, a decimal point, and a
      leading plus or minus sign, and must be right-justified and padded with
      blanks.  The numbers following the word NUMERIC specify the maximum
      number of digits which are allowed on either side of the decimal.  The
      decimal point is not assumed and must be contained in the  field in its
      correct position.  For example, the format "NUMERIC 3.2" allows 3
      digits preceding the decimal point and 2 following it (a total length
      of 6 characters).  The format "NUMERIC S3.2" allows a leading plus or
      minus sign (a total length of 7).  If a field's format description does
      not contain a decimal point,  then a decimal point is not allowed in the
      field.  If a field's format description does not contain an "S", then a
      sign is not allowed in the field.

      Explanation of NUMERIC fields in Format A.

      In the examples below the format NUMERIC 3.2 is described.

      (Quotation marks indicate limits of the field described and are not
      included in the format.)

      If the value of the field is  10.1:
      The columns  in the format will appear as:   "10.10"  (six columns).
                                   H-3                                ILM01.0

-------
      The table below demonstrates several examples:

                          Value       Appears oh Format

                            10.1             "10.10"
                          10.11             "10.11"
                         100.11            "100.11"
                             100            "100.00"
                             .29              "0.29"
                        -100.129           Invalid
                          -10.1           Invalid
      The following table presents examples of NUMERIC S3.2:

                          Value       Appears on Format

                            10.1            "  10.10" (seven columns)
                         -10.11            " -10.11"
                        -100.11            "-100.11"
                        -1000.1            Invalid
                             100            " 100.00"
                            -.22            "  -0.22"
                          -.239            "  -0.24"

2.    Record Types

2.1   Format A consists of variable length ASCII records.  The last two bytes
      of each record must contain "carriage return" and "line feed",
      respectively.  Unused bytes in partially filled fields must be blank-
      filled.

2.2   Format A has three types of records:   Header Records,  Detail Records
      and Comment Records.

          Tvoe               Tvpe ID            Contents

          Header               H               Nonrepeating fields which
                                               together are  unique to the
                                               associated hardcopy form
          Detail               D               A group  of fields that are
                                               repeated on a form, and are
                                               uniquely positioned by (e.g.,
                                               Analyte  Chemical Symbol)

          Comment              C               Nonrepeating  fields containing
                                               text that comments on informa-
                                               tion reported on the form

      The format for Comment Records  is  the  same for all forms,  and is
      described after all  other formats.

2.3   The first 5 bytes  of each record contain  the  FORM ID,  identifying the
      Inorganic Analysis Reporting  Form  for which  the record contains data.
      The ID must be left-justified in the field.
                                   H-4                                ILM01.0

-------
             FORM ID                      FORM NAME

              COVER      COVER PAGE -  INORGANIC ANALYSES  DATA PACKAGE
              I           INORGANIC  ANALYSIS DATA SHEET
              11(1)      INITIAL AND CONTINUING CALIBRATION VERIFICATION
              11(2)      CRDL STANDARD FOR AA AND ICP
              III        BLANKS
              IV          1C?  INTERFERENCE  CHECK SAMPLE
              V(l)        SPIKE SAMPLE  RECOVERY
              V(2)        POST DIGEST SPIKE SAMPLE RECOVERY
              VI          DUPLICATES
              VII        LABORATORY CONTROL SAMPLE
              VIII        STANDARD ADDITION RESULTS
              IX          ICP  SERIAL DILUTIONS
              X           INSTRUMENT DETECTION LIMITS  (QUARTERLY)
              XI          ICP  INTERELEMENT  CORRECTION  FACTORS (ANNUALLY)
              XII        ICP  LINEAR RANGES (QUARTERLY)
              XIII        PREPARATION LOG
              XIV        ANALYSIS RUN  LOG

      Following  the FORM ID  is a two-byte,  left-justified,  FORM SUFFIX, which
      must be unique for each set  of  records that correspond  to one  hardcopy
      form.   For example, records  for the first  occurrence  of a form must
      contain the  suffix AA.   Records for the  second occurrence must contain
      AB, and the  twenty-seventh occurrence  would contain BA.

      The 8th byte of each record  contains the TYPE  ID, which specifies what
      kind of data the record contains (see  paragraph 2.2).

      Records with the same  FORM ID and FORM SUFFIX  must be grouped  together
      in the  file.  Within each  FORM  ID/FORM SUFFIX  group,  there may be only
      one HEADER record,  and  it  must  come  first.  DETAIL records must follow
      the HEADER record.  The  COMMENT records, which are optional, must come
      last in the group,  and be  in sequence  corresponding to  the form.

      The FORM ID/FORM SUFFIX group(s) for the COVER PAGE(S) must come first
      in .the  file.  After the COVER PAGE(S), the FORM ID/FORM SUFFIX groups
      do not have to be in any specific order.  For example, a set of
      HEADER/DETAIL/COMMENT records for FORM V could come before records for
      FORM I, as  long as the records within the group are in the correct
      order.

3.     Record Length

    Table 3.1 summarizes  the  length (excluding carriage  return/line feed)  and
    (in  parentheses)  the  number of  records in Format  A.   The maximum number
    of detail and comment records is shown, corresponding to a submission  of
    hardcopy  foras  on which  information is written on all possible lines.
                                   H-5                                ILM01.0

-------
                           Table 3.1 Format A Summary

      Form                 	Record	
                            Header        Detail        Comment

      Cover                80a(l)b        25(20)         78(4)
      I                     90(1)         31(24)         73(4)
      11(1)                 32(1)         65(24)
      11(2)                 32(1)         66(23)
      III                   18(1)         59(24)
      IV                    32(1)         64(23)
      V(l)                  33(1)         66(24)         78(4)
      V(2)                  23(1)         62(24)         78(4)
      VI                    38(1)         56(24)
      VII                   32(1)         68(24)
      VIII                   8(1)         69(32)
      IX                    23(1)         44(23)
      X                     52(1)         29(23)         78(4)
      XI(1)                 28(1)         77(23)         78(4)
      XI(2)                 28(1)         77(23)         78(4)
      XII                   28(1)         29(23)         78(4)
      XIII                  10(1)         32(32)
      XIV                   38(1)         59(32)
      a Length of record in bytes (excluding carriage return/line feed).
        Maximum number of records required for a form.
  Record Listing

The remainder of this section contains detailed specifications for every
record required for a full set of hardcopy forzis.
                                H-6                                 ILM01.0

-------
                                  COVER PAGE
INORGANIC ANALYSES  DATA PACKAGE COVER PAGE HEADER RECORD:
COLUMN(S)

   1-5
   6-7
    8
   9-33
  34-43
  44-49
  50-54
  55-60
  61-66
  67-71
  72-74
  75-77
  78-80
 5
 2
 I
25
10
 6
 5
 6
 6
 5
 3
 3
 3
CONTENTS

1 COVER'
FORM SUFFIX
'H'
LAB NAME
CONTRACT
LAB CODE
CASE NUMBER
SAS NUMBER
SDG NUMBER
SOW NUMBER
ICP INT CORRECTIONS
ICP BG CORRECTIONS
RAW DATA BEFORE
CHARACTER
CHARACTER
CHARACTER
CHARACTER
CHARACTER
CHARACTER
CHARACTER
'YES'/'NO'
'YES'/'NO'
'YES'/'NO'/BLANK
NOTE:   The LAB NAME, CONTRACT, LAB  CODE, CASE NUMBER, SAS  NUMBER, AND  SDG
        NUMBER, which are  contained  in the   COVER PAGE HEADER record,  are  not
        repeated in the  HEADER records of  the other  forms.  Each  form's
        HEADER record contains only  data which are unique to  the  DETAIL
        records which follow it.
INORGANIC ANALYSES DATA  PACKAGE COVER PAGE  DETAIL RECORDS:

COLUMN(S)    LENGTH       CONTENTS                    FORMAT
   1-5
   6-7
    8
   9-15
  16-25
 5
 2
 1
 7
10
'COVER'
FORM SUFFIX
'D'
EPA SAMPLE NO.
LAB SAMPLE ID NO.
CHARACTER
CHARACTER
                                   H-7
                                                                       ILM01.0

-------
INORGANIC ANALYSIS DATA SHEET HEADER RECORD:
COLUMNCS)
CONTENTS
FORMAT
1-5
6-7
.8
9-15
16-20
21-30
31-33
34-41
42-46
47-51
52-60
61-69
70-75
76-31
82-87
88-90
5
2
1
7
5
10
3
8
5
5
9
9
6
6
6
3
                          'I
                          FORM SUFFIX
                          'H'
                          EPA SAMPLE NO.
                          MATRIX
                          LAB SAMPLE ID
                          LEVEL
                          DATE RECEIVED
                          PERCENT SOLIDS
                          CONCENTRATION UNITS
                          COLOR BEFORE
                          COLOR AFTER
                          CLARITY BEFORE
                          CLARITY AFTER
                         TEXTURE
                         ARTIFACTS
                           CHARACTER
                           'WATER'/'SOIL'
                           CHARACTER
                           'LOW /'MED'
                           MM/DD/YY
                           NUMERIC 3.1
                           'UG/L  '/  'MG/KG'
                           CHARACTER
                           CHARACTER
                           CHARACTER
                           CHARACTER
                           CHARACTER
                           'YES'/BLANK
INORGANIC ANALYSIS DATA SHEET DETAIL RECORDS:

COLUMN(S)    LENGTH       CONTENTS

   1-5          5         'I
   6-7          2         FORM SUFFIX
    8           1         'D'
   9-10         2         ANALYTE SYMBOL
  11-22  -      12         CONCENTRATION
   23           1         CONC FLAG (C)
  24-29         6         QUALIFIER (Q)
  30-31        2         METHOD (M)
                           FORMAT
                           CHARACTER
                           NUMERIC 9.2
                           'B'/'U'/BLANK
                           UP TO 6 ONE-CHARACTER
                           FLAGS (OTHER THAN  'B'
                           OR 'U')
                           METHOD CODE/'NR'
                                   H-8
                                             ILM01.0

-------
                               FORM  II  (PART  1)
INITIAL AND CONTINUING CALIBRATION VERIFICATION HEADER RECORD:

COLUMN(S)    LENGTH       CONTENTS                    FORMAT
                          'II(l)'
                          FORM SUFFIX
                          'H'
                          INIT CAL SOURCE             CHARACTER
                          CONT CAL SOURCE             CHARACTER
1-5
6-7
8
9-20
21-32
5
2
1
12
12
INITIAL AND CONTINUING CALIBRATION VERIFICATION DETAIL RECORDS:

COLUMN fS")   LENGTH       ONTSNTS                    70R.MA1
                         'IKD'
                         FORM SUFFIX
                         'D'
                         ANALYTE SYMBOL              CHARACTER
                         INITIAL CAL TRUE
                         INITIAL CAL FOUND
                         INITIAL CAL %R
                         CONT CAL TRUE
                         CONT CAL FOUND 1
                         CONT CAL %R 1
                         CONT CAL FOUND 2
                         CONT CAL %R 2
                         METHOD (M)
1-5
6-7
8
9-10
11-17
18-25
26-30
31-37
38-45
46-50
51-58
59-63
54-65
5
2
1
2
7
8
5
7
8
5
3
5
2
NUMERIC
NUMERIC
NUMERIC
NUMERIC
NUMERIC
NUMERIC
KUMERIC
NUMERIC 3.1
METHOD CODE/'
5.1
5.2
3.1
5.1
5.2
3.1
5.2
                                   H-9
                                                                      ILM01.0

-------
                               FORM II (PART 2)
CRDL STANDARD FOR AA AND ICP HEADER RECORD:

COLUMNrS)    LENGTH       CONTENTS
                          '11(2)'
                          FORM SUFFIX
                          .'H'
                          AA STANDARD SOURCE
                          ICP STANDARD SOURCE
1-5
6-7
8
9-20
21-32
5
2
1
12
12
                            FORMAT
                            CHARACTER
                            CHARACTER
CRDL STANDARD FOR AA AND ICP DETAIL RECORDS:
COLUMN (S")
1-5
6-7
8
9-10
11-17
18-26
27-31
32-38
39-47
48-52
53-61
62-66
5
2
1
2
7
9
5
7
9
5
9
5
CONTENTS

'11(2)'
FORM SUFFIX
'D'
ANALYTE SYMBOL
AA TRUE
AA FOUND
AA %R
ICP INIT TRUE
ICP INIT FOUND
ICP INIT %R
ICP FINAL FOUND
1C? FINAL %R
FORMAT
                                                     CHARACTER
                                                     NUMERIC 5.1
                                                     NUMERIC 6.2
                                                     NUMERIC 3.1
                                                     NUMERIC 5.1
                                                     NUMERIC 6.2
                                                     NUMERIC 3.1
                                                     NUMERIC 6.2
                                                     NUMERIC 3.1
                                   H-10
                                              ILM01.0

-------
                                   FORM III
BLANKS HEADER RECORD:

COLUMN CS')    LENGTH
1-5
6-7
8
9-13
14-18
5
2
1
5
5
             CONTENTS

             'III  '
             FORM SUFFIX
             'H'
             PREP BLANK MATRIX
             PREP BLANK UNITS
                                                     'WATER'/'SOIL '
                                                     'UG/L  '/'MG/KG'
BLANKS DETAIL RECORDS:
COLUMN(SI
LENGTH
1-5
6-7
8
9-10
11-18
19
20-27
28
29-36
37
38-45
46
47-55
57
58-59
5
2
1
2
8
1
8
1
8
1
8
1
10
1
2
CONTENTS

'III   '
FORM SUFFIX
'D'
ANALYTE SYMBOL
INITIAL CAL BLANK
INITIAL CAL FLAG (C)
CONT CAL BLANK 1
CC BLANK 1 FLAG (C)
CONT CAL BLANK 2
CC BLANK 2 FLAG (C)
CONT CAL BLANK 3
CC BLANK 3 FLAG (C)
PREPARATION BLANK
PREP BLANK FLAG (C)
METHOD (M)
                                                     CHARACTER
                                                     NUMERIC S5.1
                                                     'B'/'U'/BLANK
                                                     NUMERIC S5.1
                                                     'B'/'U'/BLANK
                                                     NUMERIC S5.1
                                                     'B'/'U'/BLANK
                                                     NUMERIC S5.1
                                                     'B'/'U'/BLANK
                                                     NUMERIC S3. 3
                                                     'B'/'U'/BLANK
                                                     METHOD CODE/'NR'
                                   H-ll
                                                                      ILM01.0

-------
                                    FORM IV


ICP INTERFERENCE CHECK SAMPLE HEADER RECORD:

COLUMNfS)    LENGTH      CONTENTS                    FORMAT

   1-5          5         'IV
   6-7          2         FORM SUFFIX
    8           1         'H'
   9-20        12         ICP ID NUMBER               CHARACTER
  21-32        12         ICS SOURCE                  CHARACTER
ICP INTERFERENCE CHECK SAMPLE DETAIL RECORDS:

COLUMN CS1)    LENGTH      CONTENTS                    FORMAT

   1-5          5         'IV
   6-7          2         FORM SUFFIX
    8           1         'D'
   9-10         2         ANALYTE S-fllBOL              CHARACTER
  11-16         6         TRUE A                      NUMERIC  6
  17-22         6         TRUE AB                     NUMERIC  6
  23-29         7         INITIAL A                   NUMERIC  S6
  30-38         9         INITIAL AB                  NUMERIC  S6.1
  39-43         5         INITIAL %R                  NUMERIC  3.1
  44-50         7         FINAL A                     NUMERIC  S6
  51-59         9         FINAL AB                    NUMERIC  S6.1
  60-64         5         FINAL %R                    NUMERIC  3.1
                                  H-12                                ILM01.0

-------
                                FORM V rPART
 SPIKE  SAMPLE RECOVERY  HEADER  RECORD:
COLUMNrs)

   1-5
   6-7
    8
   9-15
  16-20
  21-23
  24-28
  29-33
  5
  2
  1
  7
  5
  3
  5
  5
CONTENTS

'V(l)  '
FORM SUFFIX
'H'
EPA SAMPLE NO.
MATRIX
LEVEL
CONCENTRATION UNITS
SAMPLE % SOLIDS
                                       FORMAT
                            CHARACTER
                            'WATER'/'SOIL  '
                            'LOW /'MED'
                            'UG/L  '/ 'MG/KG'
                            NUMERIC 3.1
SPIKE SAMPLE RECOVERY DETAIL RECORDS:
COLUMNCS)

   1-5
   6-7
    8
   9-10
  11-16
  17-30
   31
  32-44
   45
  46-56
  57-63
   64
  65-66  '
 5
 2
 I
 2
 6
14
 1
13
 1
11
 7
 1
 2
CONTENTS

'V(l)  '
FORM SUFFIX
'D'
ANALYTE SYMBOL
CONTROL LIMIT %R
SPIKED SAMPLE RESULT
SSR FLAG (C)
SAMPLE RESULT
SR FLAG (C)
SPIKE ADDED
PERCENT RECOVERED
QUALIFIER (Q)
METHOD (M)
                                      FORMAT
                            CHARACTER
                            '75-125'/BLANK
                     NUMERIC 9.4
                            'B'/'U'/BLANK
                             NUMERIC  8.4
                            'B'/'U'/BLANK
                            NUMERIC 8.2
                            NUMERIC S4.1
                            'N'/BLANK
                            METHOD CODE/'NR'
                                  H-13
                                                        ILM01.0

-------
                                FORM V (PART 2)
POST DIGEST SPIKE SAMPLE RECOVERY HEADER RECORD:
COLUMN(SI
1-5
6-7
8
9-15
16-20
21-23
5
2
1
7
5
3
           CONTENTS

           'V(2)  '
           FORM SUFFIX
           •H'
           EPA  SAMPLE NO.
           MATRIX
           LEVEL
                            FORMAT
                                                     CHARACTER
                                                     'WATER'/'SOIL
                                                     •LOW'/'MED'
POST DIGEST SPIKE SAMPLE RECOVERY DETAIL RECORDS:
COLUMNCS)

   1-5
   6-7
    8
   9-10
  11-16
  17-28
   29
  30-41
   42
  43-52
  53-59
   60
  61-62
 5
 2
 1
 2
 6
12
 1
12
 1
10
 7
 1
 2
CONTENTS

'V(2) '
FORM SUFFIX
'D'
ANALYTE SYMBOL
CONTROL LIMIT %R
SPIKED SAMPLE RESULT
SSR FLAG (C)
SAMPLE RESULT
SR FLAG (C)
SPIKE ADDED
PERCENT RECOVERED
QUALIFIER (Q)
METHOD (M)
CHARACTER
BLANK
NUMERIC 9.2
'B'/'IJ'/KLANK
NUMERIC 9.2
•B'/'U'/BLANK
NUMERIC 8.1
NUMERIC S4.1
BLANK
METHOD CODE/'NR'
                                   H-14
                                                        ILM01.0

-------
                                    FORM VI
DUPLICATES HEADER  RECORD:
COLUMN(S)
LENGTH
1-5
6-7
8
9-15
16-20
21-23
24-28
29-33
34-38
5
2
1
7
5
3
5
5
5
 CONTENTS

 •VI
 FORM SUFFIX
 'H'
 EPA SAMPLE NO.
 MATRIX
 LEVEL
 CONCENTRATION UNITS
 SAMPLE % SOLIDS
 DUPLICATE % SOLIDS
                                                     CHARACTER
                                                     'WATER'/'SOIL '
                                                     'LOW'/'MED'
                                                     'UG/L  '/'MG/KG'
                                                     NUMERIC 3.1
                                                     NUMERIC 3.1
DUPLICATES DETAIL RECORDS:
COLUMN(S)
LENGTH
1-5
6-7-
8
9-10
11-17
18-31
32
33-46
47
48-53
54
55-56
5
2
1
2
7
14
1
14
1
6
1
2
CONTENTS

'VI   '
FORM SUFFIX
'D'
ANALYTE SYMBOL
CONTROL LIMIT
SAMPLE
SAMPLE FLAG (C)
DUPLICATE
DUPLICATE FLAG (C)
RPD
QUALIFIER (Q)
METHOD (M)
FORMAT
                                                     CHARACTER
                                                      NUMERIC 5.1
                                                     NUMERIC 9.4
                                                     'B'/'U'/BLANK
                                                     NUMERIC 9.4
                                                     'B'/'U'/BLANK
                                                     NUMERIC 4.1
                                                     '*'/BLANK
                                                     METHOD CODE/'NR'
                                   H-15
                                                                       ILM01.0

-------
                                   FORM VII


LABORATORY CONTROL SAMPLE HEADER RECORD:

COLUMNfS)    LENGTH      CONTENTS                    FORMAT

   1-5          5        'VII  '
   6-7          2        FORM SUFFIX
    8           1        'H'
   9-20        12        SOLID LCS SOURCE            CHARACTER
  21-32        12        AQUEOUS LCS SOURCE          CHARACTER
LABORATORY CONTROL SAMPLE DETAIL RECORDS:

COLUMN ("S")    LENGTH       CONTENTS                    FORMAT

   1-5          5         'VII  '
   6-7          2         FORM SUFFIX
    8           1         'D'
   9-10         2         ANALYTE SYMBOL              CHARACTER
  11-17         7         AQUEOUS TRUE                NUMERIC  5.1
  18-25         8         AQUEOUS FOUND               NUMERIC  5.2
  26-30         5         AQUEOUS % RECOVERED         NUMERIC  3.1
  31-38         8         SOLID TRUE                  NUMERIC  6.1
  39-46         8         SOLID FOUND                 NUMERIC  6.1
   47           1         SOLID FOUND FLAG  (C)         'B'/'U'/BLANK
  48-55         8         SOLID LOWER LIMIT           NUMERIC  6.1
  56-63         8         SOLID UPPER LIMIT           NUMERIC  6.1
  64-68         5         SOLID % RECOVERED           NUMERIC  3.1
                                  H-16                                ILM01.0

-------
                                  FORM VIII
STANDARD ADDITION RESULTS HEADER RECORD:

COLUMN(S)    LENGTH      CONTENTS
   1-5
   6-7
    8
                                     FORMAT
5
2
I
 'VIII  '
 FORM SUFFIX
 'H'
NOTE:   Although there are no fields which occur only once per FORM VIII,  the
        HEADER record must be included as a place holder,  indicating that
        DETAIL records follow.
STANDARD ADDITION RESULTS DETAIL RECORDS:

COLUMNCS)    LENGTH      CONTENTS
   1-5
   6-7
    8
   9-15
  16-17
  18-22
  23-28
  29-33
  34-39
  40-44
  45-50
  51-55
  56-62
  63-68
   69
5
2
1
7
2
5
6
5
6
5
6
5
7
6
1
 'VIII '
 FORM SUFFIX
 'D'
 EPA SAMPLE NO.
 ANALYTE SYMBOL
 0 ADD ABSORBANCE
 1 ADD CONCENTRATION
 1 ADD ABSORBANCE
 2 ADD CONCENTRATION
 2 ADD ABSORBANCE
 3 ADD CONCENTRATION
 3 ADD ABSORBANCE
 FINAL CONCENTRATION
CORRELATION COEF (R)
QUALIFIER (Q)
                                     FORMAT
CHARACTER
CHARACTER
NUMERIC
NUMERIC
NUMERIC
NUMERIC
NUMERIC
NUMERIC
NUMERIC 1.3
NUMERIC 5.1
NUMERIC 1.4
'+'/BLANK
1.3
3.2
1.3
3.2
1.3
3.2
                                   H-17
                                                       ILM01.0

-------
                                    FORM IX
ICP SERIAL DILUTIONS HEADER RECORD:

COLUMN(S)    LENGTH      CONTENTS
   1-5
   6-7
    8
   9-15
  16-20
  21-23
 5
 2
 1
 7
 5
 3
'IX
FORM SUFFIX
'H'
EPA SAMPLE NO.
MATRIX
LEVEL
                                      FORMAT
CHARACTER
'WATER'/'SOIL
'LOW/'MED'
ICP SERIAL DILUTIONS DETAIL RECORDS:

COLUMN(S)    LENGTH       CONTENTS
   1-5
   6-7
    8
   9-10
  11-22
   23
  24-35
   36
  37-41
   42
  43-44
 5
 2
 1
 2
12
 1
12
 1
 5
 1
 2
'IX
FORM SUFFIX
'D'
ANALYZE STCIBOL
INIT SAMPLE (I)
INIT SAMPLE FLAG (C)
SERIAL DILUTION (S)
DILUTION FLAG (C)
PERCENT DIFFERENCE
QUALIFIER (Q)
METHOD (M)
                                      FORMAT
CHARACTER
NUMERIC 9.2
'B'/'U'/BLANK
NUMERIC 9.2
'B'/'U'/BLANK
NUMERIC 3.1
'E'/BLANK
METHOD CODE/'NR'
                                   H-18
                                                        ILM01.0

-------
                                    FORM X
 INSTRUMENT DETECTION LIMITS  (QUARTERLY) HEADER RECORD:

 COLUMNCS)    LENGTH       CONTENTS                    FOE
   1-5
   6-7
    8
   9-16
  17-28
  29-40
  41-52
 5
 2
 1
 8
12
12
12
 FORM  SUFFIX
 'H'
 DATE
 ICP ID NUMBER
 FLAME AA ID NUMBER
 FURNACE AA ID NUMBER
MM/DD/YY
CHARACTER
CHARACTER
CHARACTER
INSTRUMENT DETECTION LIMITS (QUARTERLY) DETAIL RECORDS:

COLUMNCS)    LENGTH      CONTENTS                    FORMAT
   1-5
   6-7
    8
   9-10
  11-17
  18-19
  20-27
  28-29
 2
 1
 2
 7
 2
 8
 2
'X
FORK SUFFIX
'D'
ANALYTE SYMBOL
WAVELENGTH
BACKGROUND
IDL
METHOD (M)
CHARACTER
NUMERIC 4.2
'BS'/'BD'/'BZ'/BLANK
NUMERIC 6.1
METHOD CODE/'NR'
                                  H-19
                                                                       ILM01.0

-------
                                FORM XI  (PART 1)
 ICP  INTERELEMENT  CORRECTION FACTORS  (ANNUALLY)  HEADER  RECORD:

 COLUMN(S)    LENGTH       CONTENTS                    FORMAT
    1-5
    6-7
    8
    9-20
  21-28
 5
 2
 1
12
 8
 'XI(l)'
 FORM SUFFIX
 'H'
 ICP ID NUMBER
 DATE
                  CHARACTER
                  MM/DD/YY
ICP INTERELEMENT CORRECTION FACTORS (ANNUALLY) DETAIL RECORDS:

COLUMN(S)    LENGTH      CONTENTS                    FORMAT
   1-5
   6-7
    8
   9-10
  11-17
  18-19
  20-29
  30-31
  32-41
  42-43
  44-53
  54-55
  56-65
  66-67
  68-77
 5
 2
 1
 2
 7
 2
10
 2
10
 2
10
 2
10
 2
10
'XI(l)'
FORM SUFFIX
'D'
ANALYTE SYMBOL
WAVELENGTH
ELEMENT
ELEMENT
SYMBOL
FACTOR
ELEMENT 2 SYMBOL
ELEMENT 2 FACTOR
ELEMENT 3 SYMBOL
ELEMENT 3 FACTOR
ELEMENT 4 SYMBOL
ELEMENT 4 FACTOR
ELEMENT 5 SYMBOL
ELEMENT 5 FACTOR
CHAKACTE?.
NUMERIC 4.2
AL
NUMERIC SI.7
CA
NUMERIC SI. 7
FE
NUMERIC SI. 7
MG
NUMERIC SI. 7
CHARACTER
NUMERIC SI.7
NOTE:    ELEMENTS 1, 2, 3 and 4 SYMBOL can only be AL, CA, FE and MG
        respectively.   ELEMENT 5 Symbol can be any other analyte symbol.
                                   H-20
                                                        ILM01.0

-------
                               FORM XI (PART 2)
ICP INTERELEMENT CORRECTION FACTORS (ANNUALLY) HEADER RECORD:

COLUMN(S)    LENGTH      CONTENTS                    FORMAT
                           CHARACTER
                           MM/DD/YY
1-5
6-7
8
9-20
21-28
5
2
1
12
8
                         FORM SUFFIX
                         'H'
                         ICP ID NUMBER
                         DATE
                         FORM SUFFIX
                         'D'
                         ANALYTE SYMBOL
                         WAVELENGTH
ICP INTERELEMENT CORRECTION FACTORS (ANNUALLY) DETAIL RECORDS:

COLUMNSS)    LENGTH      CONTENTS                    FORMAT
CHARACTER
NUMERIC 4.2
CHARACTER
NUMERIC SI.7
CHARACTER
NUMERIC SI.7
CHARACTER
NUMERIC SI.7
CHARACTER
NUMERIC SI.7
CHARACTER
NUMERIC SI. 7
1-5
6-7
8
9-10
11-17
18-19
20-29
30-31
32-41
42-43
44-53
54-55
56-65
66-67
68-77
5
2
1
2
7
2
10
2
10
2
10
2
10
2
10
                         ELEMENT 1
                         ELEMENT 1
                         ELEMENT 2
                         ELEMENT 2
                         ELEMENT 3
                         ELEMENT 3
          SYMBOL
          FACTOR
          SYMBOL
          FACTOR
          SYMBOL
          FACTOR
ELEMENT 4 SYMBOL
ELEMENT 4 FACTOR
        5
        5
ELEMENT
ELEMENT
                                   SYMBOL
                                   FACTOR
                                   H-21
                                                                       ILM01.0

-------
                                   FORM XII
TCP LINEAR RANGES (QUARTERLY) HEADER RECORD:

COLUMN(S)    LENGTH       CONTENTS
   1-5
   6-7
    8
   9-20
  21-28
 5
 2
 1
12
 8
'XII  '
FORM SUFFIX
'H'
ICP ID NUMBER
DATE
                                      FORMAT
CHARACTER
MM/DD/YY
ICP LINEAR RANGES (QUARTERLY) DETAIL RECORDS:

COLUMN(S)    LENGTH       CONTENTS
                                      FORMAT
   1-5
   6-7
    8
   9-10
  11-16
  17-27
  28-29
 5
 2

 2
 6
11
 2
'XII  '
FORM SUFFIX
'D'
ANALYTE SYMBOL
INTEGRATION TIME
CONCENTRATION
METHOD (M)
(ICP IS ASSUMED
IF BLANK)
CHARACTER
NUMERIC 3.2  (SECONDS)
NUMERIC 9.1
'NR'/BLANK
                                   H-22
                                                        ILM01.0

-------
                                    FORM  XIII


PREPARATION LOG HEADER RECORD:

COLUMNCS)    LENGTH       CONTENTS                    FORMAT

  1-5           5         'XIII '
  6-7           2         FORM SUFFIX
    8           1         'H'
 9-10           2         METHOD                      METHOD CODE
PREPARATION LOG DETAIL RECORDS:

COLUMNCS)    LENGTH    .   CONTENTS                     FORMAT

   1-5          5         'XIII  '
   6-7          2         FORM SUFFIX
    8           :.         ' D •
   9-15         7         EPA  SA11PLE  ir.J:.'3ER           CHARACTER
  16-23         '          PREP DATE                    MM/DD/YY
  24-28         5'        WEIGHT                      NUMERIC  2.2
  29-32         4         VOLUME                      NUMERIC  4
                                   H-23                                 ILM01.0

-------
                                    FORM XIV
ANALYSIS RUN LOG HEADER RECORD:

COLUMN(S)    LENGTH      CONTENTS
1-5
6-7
8
9-20
21-22
23-30
31-38
5
2
1
12
2
8
8
                          'XIV   '
                          FORM  SUFFIX
                          'H'
                          INSTRUMENT ID NUMBER
                          METHOD
                          START DATE
                          END DATE
                            CHARACTER
                            METHOD CODE
                            MM/DD/YY
                            MM/DD/YY
ANALYSIS RUN LOG DETAIL RECORDS:
COLUMN(S)
CONTENTS
FORMAT
1-5
6-7
8
9-16
17-24
25-28
29-35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
5
2
1
8
8
4
7
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
                         ' XIV  '
                         FORM SUFFIX
                         'D'
                         EPA SAMPLE NUMBER
                         DILUTION FACTOR
                         TIME
                         PERCENT RECOVERED
                         ANALYTE (AL)
                         ANALYTE (SB)
                         ANALYTE (AS)
                         ANALYTE (BA)
                         ANALYTE (BE)
                         ANALYTE (CD)
                         ANALYTE (CA)
                         ANALYTE (CR)
                         ANALYTE (CO)
                         ANALYTE (CU)
                         ANALYTE (FE)
                         ANALYTE (PB)
                         ANALYTE (MG)
                         ANALYTE (MN)
                         ANALYTE (HG)
                         ANALYTE (NI)
                         ANALYTE (K)
                         ANALYTE (SE)
                         ANALYTE (AG)
                         ANALYTE (NA)
                         AiN'ALYTE (TL)
                         ANALYTE (V)
                         ANALYTE (ZN)
                         ANALYTE (CN)
                            CHARACTER
                            NUMERIC  5.2
                            HHMM
                            NUMERIC  S4.1
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X-/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "XVBLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X"/BLANK
                            "X-'/BLANK
                            "X"/BLANK
                            "X"/BLANK
                              "/BLANK
                            X" /BLANK
                            "X"/BLANK
"X
ii V"
                                   H-24
                                                                       ILM01.0

-------
                                COMMENT RECORDS
COMMENT records are optional for any FORM ID/FORM SUFFIX group. There may be
up to 4 COMMENT records per group.  They must come after the DETAIL records
and be formatted as follows:
COLUMN(S)    LENGTH      CONTENTS                    FORMAT '

   1-5          5        FORM ID
   6-7          2        FORM SUFFIX
    8           1        'C'
   9-78        70        COMMENTS                    FREE FORM TEXT
The text may be in paragraph form if desired, but must be contained in columns
9 through 78 only.  The key fields must be repeated in columns 1 through 8 of
each line.
                                   H-25                                ILM01.0

-------
                                   SECTION III
                             FORMAT B SPECIFICATION
1.    Introduction

1.1   This constitutes  the  implementation of the EPA standard for media and
      record formats to be  used in transmission of analytical results for the
      CLP inorganics program.  The following points should be noted:

      1.1.1    The column border  "|"  is  placed  between  fields  to permit  these
               records  to be prepared by programs written for  laboratory and
               quality  assurance automation  systems, and  the detection of
               possible field shift.  They also assist  in visual clarity.

      1.1.2    Record formats contain sequence  numbers  and checksums  to  be
               consistent with requirements  for a. future  error-free
               telecommunications format.

2.    Record Types

2.1   There are four groups of record types in the  reporting format, as shcvr.
      below.  Detailed  record formats follow.

Type    Name                Contents

10      Run Header          Contains information pertinent to the
                            whole production run.   See production
                            run definition below.

20      Sample Header       Contains sample-identifying information,
                            or corresponding information for calibra-
                            tions,  QC samples,  instrument performance
                            checks, etc.

30      Results Record      Contains any final result on a sample,
                            calibration or QC sample and identifying
                            information.

90      Comments Record    Contains free-form comments and/or  other
                            miscellaneous information.

2.2   All record types  given are mandatory.   Type,  10 representing the
      analytical run, contains the instrument and run IDs which act as an
      identifying label for the run.   All 10, 20,  30,  and 90 series records
      following that record pertain to the same analytical run.  Type 20,
      representing the  sample, contains the EPA Sample ID which acts as an
      Identifying label for the sample.  The QC code indicates  whether the data
      is from an environmental sample,  calibration, or QC sample.  All 20, 20,
      and 90 series records following that record pertain to the same sample.
      Type 30, representing an individual analyte,  contains an  indentifier to
                                        H-26                        ILM01.0

-------
       identify the analyte.   All 30 series  records  following that record
       pertain to the same analyte.

       Production Runs

       A production run represents a "group"  or  "batch"  of samples that are
       processed in a continuous  sequence under  relatively stable  conditions.
       Specifically:

       Calibration -  All samples  in a run use the  same initial ^calibration data.

       Method  -  Cons tant.
       Instrument conditions  -  Constant  throughout a run.  Results  obtained on
       different instruments  cannot be combined  in one  run.

       Thus, each separate  group  of analyses on  each instrument will  consist of
       a  separate production  run,  and must be  reported  in a separate  file.

4.     Record  Sequence

4.1    A  Run Header  (type 10) record must be present as  the first record in the
       file.   Further occurrences  of the type  10 record  in the file are  not
       allowed..

       A  type  16  record must  immediately follow  the type 10 record.   Further
       occurrences of the type  15  record in the  file are not allowed.

       A  type  20  record must  immediately follow  the type 16 record  as  a  header
       for the calculated run-wide instrument parameters (the quarterly  and
       annual  instrument parameters) .  This is the only occurrence  of type  20
       record  that does not correspond to an actual analysis in the run.
       Therefore,  The only fields  that are not blank in this occurrence  of  type
       20 record  are the RECORD TYPE ("20"), EPA SAMPLE NUMBER ("SIDICF"),  and
       WAVELENGTH COUNT.

       A r.inimun!  of cne group of  type 32, 34, and 35 rec~rcc —:~t irj?.3ci.c.tely
       follow  the type 16 record.  Each group consists of a type 32 record
       immediately followed by a  type 34 record  immediately followed by  a type
       35 record.  The information in each group must pertain to one  and only
       one analyte's wavelength.  The number of groups must be equivalent to the
       WAVELENGTH COUNT value and the number of wavelengths used for  analysis in
       the run.   The last group must immediately be followed by the first type
       20 record  which corresponds to an actual analysis of an instrument
       calibration standard.  After the appearance of the second type  20 record
       in the file, further occurrences of the type 32, 34, and 35  records  in
       that file  are not allowed.

4.2    Each environmental sample, calibration,  or quality control sample is
       represented by a group composed of a type 20,  21, 22,  28 records, which
      holds sample level identifying information,  followed by a ninlir.um of  one
      group composed of type 30, 31,  and 33 records for each analyte's
      wavelength.  The type 20 record holds a count for the number of analyte
      wavelengths being used to determine results.   The WAVELENGTH COUNTER  must


                                        H-27                        ILM01.0

-------
      have a value equivalent to the number of type 30 groups associated with
      each type 20 record.

      Except for the first type 20 record, all type 20 records should occur in
      the order of sample analysis.  Excluding the first type 20 record, the
      number of type 20 records in a file (run) must be equivalent to the
      number of entries reported on Form XIV for that run.

4.3   Type 90 comment records may be defined to occupy any position except
      before the type 10 (header) record.  Comments pertaining to the whole run
      such as ones on Cover Page must appear before the first type 20 record.
      Comments pertaining to a particular sample such as ones on Form i must
      appear after the type 20 record for that sample, but before the first
      type 30 record associated with that sample.  Comments pertaining to a
      particular analyte or wavelength must appear after the type 30 record of
      that wavelength, but before the type 30 record of the following
      wavelength.

5.    File/Record Integrity

      All record types must contain the following check fields to ens'ure file
      and record integrity:

Record       Field
Positior.     Length     Contents                      Remarks

 1-2          2         Record type or identifier     "10" or as appropriate

72-74         3         Record sequence number        000-999, repeated as
                        within file                   necessary
                                                                             .ju
75-78         4         Record checksum               Four hexadecimal digits ;
                                                      Calculation algorithm to
                                                      be supplied

79-80         2         Contains CR and LF

6.    Dates and Times

      Wherever a date or time-of-day is required, the information consists of
      successive groups of two right justified decimal digits each,  separated
      by "|".  Dates are given in the order YY MM DD, and times as HH MM.  All
      hours will be given as 0 to 23 using a 24 hour clock and will be local
      time.  Since some computers generating the date and time sequence may
      have difficulty producing leading zeros, these will not be required.

7.    Field format listed as CHARACTER may contain any standard ASCII character
      other than "|".  It must be left justified and padded to the right with
      blanks.  Field format listed as NUMERIC may contain numeric digits, a
      decimal point, and a leading plus or minus sign.  It must be right
      justified and padded to the left with blanks.  Except where specified
      otherwise, the numeric field must contain the minimum significance
      specified in the forms instruction of Exhibit B.  If more significance is
      used, it must be applied uniformly.  Exponent fields are numeric fields


                                        H-28                        ILM01.0

-------
      that can contain only digits and plus or minus sign.  No decimal is
      allowed.  For field formats that are specified as a choice of stings or
      characters, only those choices shown may be used.

8.    Multiple Volume Data

      There is no requirement under this format that all the data from an
      entire sample delivery group fit onto a single diskette.  However, each
      single production run must fit onto a single diskette if possible.
      However, if that is not possible, then it is necessary that all files
      start with a. type 10 record, and that the multiple type 10 records for
      each file of the same production run be identical.  If it is necessary to
      split the data from a single sample onto multiple diskettes, then the
      type 20 and following type records for that sample must be repeated.  In
      this situation, it is mandatory that columns 7-30, which collectively
      identify the sample, be identical in each diskette.

8.    Record Listing

      The remainder of this section contains detailed specifications for every
      record required for a full run.
   The checksum is defined to be the sum of the thirty-five (35) INTEGERS that
make up the data in columns 1 to 70 when the data is represented in the format
35A2 on processors which store the data bytes in left to right order.  The sum io
taken module 65536 (2  ) and represented as four (4) hexadecimal digits.  For
processors which use an A70 character representation of the data, the checksum is
the sum of all the even character position values plus 256 times the sum of all
odd character position values.
                                        H-29                        ILM01.0

-------
         FORMAT OF THE PRODUCTION  RUN FIRST HEADER RECORD  (TYPE
COLUMN'S
LENGTH
1
3
4
6
7
9
10
12
13
15
16
18
19
- 2
- 3
- 5
— 6
- 8
- 9
- 11
- 12
- 14
- 15
- 17
- 18
- 20
2
1
2
1
2
1
2
1
2
1
2
1
2
21
22
"J T
-^ —
32
35
36
42
43
58
59
69
70
71
72
75
- 21
- 30
- 31
- 34
- 35
- 41
- 42
- 57
- 58
- 68
- 69
- 70
- 71
- 74
- 78
1
9
T_
^
1
6
1
15
1
10
1
1
1
3
4
        CONTENTS


RECORD  TYPE
M I II

ANALYSIS  START YEAR
n I it

ANALYSIS  START MONTH
n | n

ANALYSIS  START DAY
ii I n

ANALYSIS  START HOUR
ti I it

ANALYSIS  START MINUTE
n I n

METHOD
                    n I it

                    BLANK
                   LAB CODS
                   it | ii

                   BLANK
                   it I ti

                   INSTRUMENT  ID
                   n | n

                   BLANK
                   it I n

                   RECORD SEQUENCE NUMBER
                   RECORD CHECKSUM
 FORMAT


"10"


YY


MM


DD


HH


MM
                                                       "AM"/"CV"/"AV"/"C"/
                                                       »CA"/"AS"/"T"
                                            CliASACTER
                                            CHARACTER
                                            NUMERIC
                                            NUMERIC
                                     H-30
                                                              ILM01.0

-------
        FORMAT OF THE PRODUCTION  RUN  SECOND HEADER RECORD (TYPE
COLUMNS
LENGTH
1
3
4
6
7
9
10
12
13
15
16
18
19
21
22
23
24
25
26
27
28
53
54
,'1
72
75
- 2
- 3
- 5
- 6
- 8
- 9
- 11
- 12
- 14
- 15
- 17
- 18
- 20
- 21
- 22
- 23
- 24
- 25
- 25
- 27
- 52
- 53
- 70
- 71
- 74
- 78
2
1
2
1
2
1
2
1
2
1
2
1
2
1
1
1
1
1
-
-L
25
1
17
1
3
4
        CONTENTS


RECORD  TYPE
it I n

ANALYSIS END YEAR
it I N

ANALYSIS END MONTH
it I n

ANALYSIS END DAY
n I n

ANALYSIS END HOUR
n I n

ANALYSIS END MINUTE
n I it

AUTO-SAMPLER USED
n I n

INTERELEMENT CORRECTIONS APPLIED
n | n

BACKGROUND CORRECTIONS APPLIED
n I n

RAW DATA GENERATED
ii I n

LABORATORY NIu-IE
n I n

BLANK
ii I n

RECORD SEQUENCE NUMBER
RECORD CHECKSUM
 FORMAT


"16"


YY


MM


DD


HH


MM


"Y"/BLANK
                                                     NUMERIC
                                                     NUMERIC
 Enter  "Y"  if an auto-sampler is used with equal analysis time and
 intervals  between analysis.

 These  are  the answers to the first two questions en the Cover Pacre.   "Y"
 equals "YES",  and "N" equals "NO".

 This is the  answer to the third question on the Cover Page.   "B"  equals
 "YES",  and "A"  equals BLANK.
                                    H-31
                                                            ILM01.0

-------
               FORMAT OF THE SAMPLE HEADER DATA RECORD  fTYPE 20)
 COLUMNS   LENGTH          CONTENTS                     FORMAT

  1-2      2      RECORD TYPE                       "20"
  3-3      1      "|"
  4-5      2      BLANK
  6-6      1      "|"
  7-14      8      EPA  SAMPLE  NUMBER                 CHARACTER
 15-15      1      "| "
 16-16      1      MATRIX                            iiniiii  2
 17-17     1      "| "
 18-24     7      BLANK
 25-25     1      "| "
 26-31     6      CASE  NUMBER                       CHARACTER
 32-32     1      "|"
 33-37     5      BLANK
 38-38     1      "|"
 39-40     2      ANALYSIS YEAR                     YY
 41 - 41     1      "|"
 42-43     2      ANALYSIS MONTH                    MM
 44-^4     1      "|"
 -'.5 - -'.6     ;•      ANALYSIS T .Y                       DD
 *-r / ~~ *± I     __        I
 48-49     :      ANALYSIS KOUR                     HH
 50-50     1      "|"
 51-52     2      ANALYSIS MINUTE                   MM
 53-53     1      "|"
 54-54     1      BLANK
 55-55     1      "|«
 56-56     1      SAMPLE UNIT CODE                  "G"/"M"  3
 57-57     1      "|"
 58-65     8      SAMPLE SIZE                       NUMERIC
                    (WET  WEIGHT OR INITIAL VOLUME)
 66-66     I      "\"
 67-69     2      ANALYTS VCAVELZirGTI-I COUNT          NUMERIC
 70-70     1      "|"
 71-71     1      "|"
 72-74     3      RECORD SEQUENCE NUMBER            NUMERIC
 75-78     4      RECORD CHECKSUM                   NUMERIC


1  EPA Sample Number as appears on Form XIV except for the  first type 20
   record.  The  first  type 20  record must have an EPA Sample Number of
   "SIDICF"

2  "1" equals "WATER", and "F" equals "SOIL"

3  "G" equals grams, and  "M" equals mL.
                                      H-32                     ILM01.0

-------
               FORMAT OF THE SAMPLE HEADER DATA RECORD  (TYPE 21)


 COLUMNS    LENGTH          CONTENTS                    FORMAT

  1-2      2      RECORD TYPE                        "21"
  3-3      1      " | "
  4-4      1      BLANK
  5-5      1      "|"
  6-6      1      CONCENTRATION LEVEL                "M"/"L"  ^
  7-7      1      " | "
  8-16      9      BLANK
 17-17      1      " | " '
 18-23      6      SAS NUMBER                        CHARACTER
 24-24      1      "|"
 25-34     10      LAB SAMPLE ID                     CHARACTER
 35-35      1      "|"
 36-36      1      "|"
 37-38      2      PREPARATION YEAR                  YY
 39-39      1      "|"
 40-41      2      PREPARATION MONTH                 MM
 42-42      1      "|"
 43-44      2      PREPARATION DAY                   DD
 • 5-45      :       <•{<••
 -.6-46      1      BLANK
 -=7 - 47      I      "|"
 48-49      2      LAB RECEIPT YEAR                  YY
 50-50      1      "|"
 51-52      2      LAB RECEIPT MONTH                 MM
 53 - 53      1      "| "
 54-55      2      LAB RECEIPT DAY                   DD
 56-56      1      "|»
 57 - 68     12      SOLUTION SOURCE                   CHARACTER
 69-69      1      "| "
 70-70      1      BLANK                             NUMERIC
 71-71      1      "| »
 ~2 - 74      3       RECORD SEQUENCE NUMBER            NUMERIC
 75-78      4       RECORD CHECKSUM                   NUMERIC


1  "M" equals  "MEDIUM",  and "L" equals "LOW".
                                      H-33                    ILM01.0

-------
              FORMAT OF THE SAMPLE HEADER DATA RECORD  fTYPE 221
COLUMNS

 1-2
 3-3
 4-40
41 - 41
42 - 46
47 - 47
48 - 55
56 - 56
57 - 64
65 - 65
66 - 70

71 - 71
72 - 74
75 - 78
 2
 1
37
 1
 5
 1
 8
 1
 8
 1
 1
 3
 4
        CONTENTS

RECORD  TYPE
n I it
BLANK
n I ii
FINAL VOLUME  IN ML
it | n
DILUTION FACTOR
n I n
BLANK
II I Tl
PERCENT SOLIDS
RECORD SEQUENCE NUMBER
RECORD CHECKSUM
 FORMAT

"22"



NUMERIC

NUMERIC
NUMERIC
(to one decimal place)

NUMERIC
NUMERIC
                                     H-34
                                                             ILM01.0

-------
             FORMAT OF THE  SAMPLE HEADER DATA RECORD (TYPE 28)


COLUMNS   LENGTH          CONTENTS                    FORMAT

 1-2     2      RECORD TYPE                        "28"
 3-3     1      " | "
 4-13    10      CONTRACT NUMBER                   CHARACTER
14-14     1      "| "
15-19     5      SOW NUMBER                         CHARACTER
20-20     1      "|"
21-26     6      SDG NUMBER                         CHARACTER
27-27     1      »|"
28-29     2      PREPARATION START HOUR            HH 1
30-30     1      "|"
31 - 70    40      BLANK
71 - 71     1      «|"
72-74     3      RECORD SEQUENCE NUMBER            NUMERIC
75-78     4      RECORD CHECKSUM                   NUMERIC


  This is the hour at  which the preparation is started.   It is  used  to
  differentiate between different  batches on  the same
                                     H-35                     ILM01.0

-------
              FORMAT OF THE  SAMPLE HEADER  DATA  RECORD (TYPE 301
 COLUMNS
  1
  3
  4
  5
  6
 15
 16
 25
 26
 31
 32
    2
    3
    4
    5
   14
   15
   24
   25
   30
   31
   34
LENGTH

  2
  1
  1
  1
  9
  1
  9
  1
  5
  1
  3
35
36
42
43
46
47
48
49
55
56
59
60
61
62
67
68
71
72
75
- 35
- 41
- 42
- 45
- 46
- 47
- 48
- 54
- 55
- 58
- 59
- 60
- 61
- 66
- 67
- 70
- 71
- 74
- 78
1
6
1
3
1
1
1
6
1
3
1
1
1
5
1
3
1
3
4
       CONTENTS

RECORD TYPE
ii I it
ANALYTE IDENTIFIER TYPE
ii I ii
ANALYTE CAS NUMBER
ii 1«
BLANK
If I H
UNITS
ii I if
CONCENTRATION QUALIFIER
                    CONCENTRATION

                    EXPONENT
                    11 j ii
                    VALUE  DESCRIPTOR
                    ti I ti
                    AMOUNT ADDED OR TRUE VALU£

                    EXPONENT
                    H I ti
                    QC VALUE  DESCRIPTOR
                    n | ii
                    QC VALUE

                    EXPONENT
                    it I n
                    RECORD SEQUENCE NUMBER
                    RECORD CKECKSUil
 FORMAT

"30"
CN FOR CYANIDE
"UG/L"/"MG/KG"

"BDL"/"LTC"/"FQC"/
»GTL"/"NAR"/"RIN"/
"REX "/BLANK 2

NUMERIC
"EVBIANK
NUMERIC
                                                    "T"/

                                                    NUMERIC
                                                    "E"/BLANK
                                                    NUMERIC
                                                    NUMERIC
                                                    "E"/BLANK
                                                    NUMERIC

                                                    NUMERIC
                                                    NUI12RIC
1

2
"C" is used  for all  analytes except cyanide.   "I" is used for cyanide.
"BDL" means  below detection limit
"NSQ" means  there is not sufficient quantity  to analyze sample according
to the protocol.
"NAI" not analyzed due  to interference    "NAR" no analysis result
required.  This is only used for the first, second,  and third addition of
MSA.  The zero addition must contain the final sample result in ug/L or
mg/Kg, as appropriate,  whether the final result is reported on Form I or
not.
Note that there is no absolute equivalent to  the final concentration on
Form VIII in Format  B.
                                      H-36
                                                           ILM01.0

-------
   "LTC" means less than the CRDL but greater than or equal to the IDL.
      "FQC" means failed quality control criteria.
   "GTL" means greater than the linear range.

   "RIN" means that the analysis result were not used to report data in the
   SDG.  The results are reported from a later reanalysis of the same sample
   aliquot.

   "REX" means that the analysis result were not used to report data in the
   SDG.  The results are reported from a later reanalysis of a repreparation
   of same sample

Note that, except for "NAR", none of these codes relief the contractor form
reporting a valid result.  They only explain why or if the result is
qualified.

3  HTH stands for a true value of the solution.  This includes the
   concentration of all (ICP as well) instrument calibration standards.  "F"
   stands for an added concentration to a sample such as a pre or post
   digestion spike, or MSA additions.

4  "P" equals percent, "C" equals correlation coefficient, and "L" equals
   control limit
                                     H-37                    ILMOl.O

-------
              FORMAT OF THE SAMPLE HEADER DATA RECORD (TYPE 31)
 COLUMNS
  1  -
  3  -
  4  -
  5  -
  6  -
2
3
4
5
6
 7-7
 8-15

15 - 15
16 - 17
18 - 18
19 - 28
29 - 29
30 - 37
38 - 38
39 - 48
49 - 49
50 - 57
58 - 53
59 - o G
69 - 69
70 - 70
71 - 71
72 - 74
75 - 78
LENGTH

  2
  1
  1
  1
  1

  1
  8

  1
  2
  1
 10
  1
  8
  1
 10
  1
  8
  1
 10
  1
  1
  1
  3
  4
       CONTENTS

RECORD TYPE
ti I n
TYPE OF DATA
it | it
TYPE OF VALUE
             ANALYTE WAVELENGTH

             n I ii
             BLANK
             it I ii
             FIRST INSTRUMENT VALUE
             ii I it
             BLANK
             if I n
             SECOND INSTRUMENT VALUE
             it g n
             BLANK
             tr | n
             THIRD INSTRUinNT VALUE
             » I ii
             BLANK
             n | ii
             RECORD SEQUENCE NUMBER
             RECORD CHECKSUM
 FORMAT

"31"

"W"

"Cn/nUn/n'B"/nF"
                                  NUMERIC
                                  (to two decimal  places)
                                  NUMERIC
                                  NUMERIC
                                  NUMERIC 2
                                  NUMERIC
                                  NUMERIC
  "C" equals concentration in ug/L,  "T" equals concentration in ug/250mL,
  "F" equals concentration in ug/50  mL, "B" equals  absorbarice,  "I" equals
  intensity, and "A" equals peak area in cm2.
  This is used to report replicate injection or exposures.   If  a single
  instrument measurement is used, then enter it in  the first instrument.
  value field, and leave the second  and third empty.   If duplicate
  instrument measurements are used,  then enter them in the  first and second
  instrument value fields in the order of their analysis, and leave the
  third field empty.  If triplicate  instrument measurements were taken, then
  enter the values in the order of their analysis.
                                     H-38
                                                       ILM01.0

-------
              FORMAT OF THE SAMPLE HEADER DATA RECORD  (TYPE 32)
COLUMNS   LENGTH           CONTENTS                    FORMAT

 1-2     2      RECORD TYPE                       "32"
 3-3     1      " | "
 4-7     4      BLANK
 8-8     1      " | "
 9-10     2      INTEGRATION TIME CODE             "IT"
11-11     1      "|"
12-17     6      INTEGRATION TIME IN SECONDS       NUMERIC
18-18     1                                         "E"/BLANK
19-21     3      EXPONENT                           NUMERIC
22-22     1      «|"
23 - 70    48      BLANK
71-71     1      »|"
72-74     3      RECORD SEQUENCE NUMBER            NUMERIC
75-78     4      RECORD CHECKSUM                   NUMERIC
                                     H-39                     ILM01.0

-------
             FORMAT OF
COLUMNS
        LENGTH
1
3
4
13
14
15
16
26
27
28
29
35
36
39
40
41
42
'48
f. 0
52
53
59
60
63
64
71
72
75
- 2
- 3
- 12
- 13
- 14
- 15
- 25
- 26
- 27
- 28
- 34
- 35
- 38
- 39
- 40
- 41
- 47
- 48
- 51
- 52
- 58
- 59
- 62
- 63
- 70
- 71
- 74
- 78
2
1
9
1
1
1
10
1
1
1
6
1
3
1
1
1
6
1
3
I
6
1
3
1
7
1
3
4
       CONTENTS


RECORD TYPE
it I tt

ANALYTE NAME
it | ii

RAW DATA AVERAGE QUALIFIER
ii I (i

RAW DATA AVERAGE
n i ft

RAW DATA %RSD QUALIFIER
N I it

RAW DATA %RSD


EXPONENT
ii j »

QC LIMIT QUALIFIER
n i ti

QC LOWER LIMIT


EXPONENT
.,,.<

QC UPPER LIMIT


EXPONENT
n i ii

CRDL IN UG/L
ii i it

RECORD SEQUENCE NUMBER
RECORD CHECKSUM
 FORMAT


"33"


CHARACTER
                                                                   JL
                                                      NUMERIC


                                                      "M"/BLAJTK


                                                      NUMERIC
                                                      "E"/BLANK
                                                      NUMERIC
                                                      NUMERIC
                                                      "E"/BLANK
                                                      NUMERIC


                                                      NUMERIC
                                                      "E"/BLANK
                                                      NUMERIC


                                                      NUMERIC


                                                      NUMERIC
                                                      NUMERIC
                                                                         "  ''
"U" means less than the  IDL,  "B" means less the CRDL and greater than or
-qual to the IDL,  "L" means  greater than

"S" flag is not applicable for Format B.
                                             -ie   nger rane.
                                     H-40
                                                            ILM01.0

-------
             FORMAT OF THE SAMPLE HEADER DATA RECORD  fTYPE
COLUMNS
LENGTH
1
3
4
14
15
23
24
30
31
34
35
41
42
45
46
60
61
/"* t1
64
66
67
69
70
71
72
75
- 2
- 3
- 13
- 14
- 22
- 23
- 29
- 30
- 33
- 34
- 40
- 41
- 44
- 45
- 59
- 60
- 62
- 63
- 65
- 66
- 68
- 69
- 70
- 71
- 74
- 78
2
1
10
1
8
1
6
1
3
1
6
1
3
1
14
1
2
-
2
1
2
1
1
1
3
4
CONTENTS
                   RECORD TYPE
                   n I n

                   BLANK
                   ii j ti

                   ANALYTE WAVELENGTH
                   n I it

                   IDL
                   n I n

                   BLANK
                   n I n

                   LINEAR RANGE
                   EXPONENT
                   n I n

                   BLANK
                   ii I n

                   YEAR COMPUTED
                   MONTH COMPUTED
                   n I n

                   DAY COMPUTED
                   ii I ii

                   BLANK
                   n I n

                   RECORD SEQUENCE NUMBER
                   RECORD CHECKSUM
 FORMAT


"34"
                                           NUMERIC
                                           (to two decimal places)


                                           NUMERIC 1
                                           NUMERIC
                                           "E"/BLANK
                                           NUMERIC
                                           NUMERIC
                                           NUMERIC
  The IDL must be a whole number for all analytes except for mercury.
  Mercury must be reported to one decimal place.
                                    H-41
                                                             ILM01.0

-------
             FORMAT OF THE SAMPLE HEADER DATA RECORD (TYPE 35)
COLUMNS
1
3
4
7
8
10
11
13
14
16
17
19
20
22
23
32
33
42
48
49
52
53
71
72
75
- 2
- 3
- 6
- 7
- 9
- 10
- 12
- 13
- 15
- 16
- 18
- 19
- 21
- 22
- 31
- 32
- 40
- 47
'- 43
- 51
- 52
- 70
- 71
- 74
- 78
LENGTH

  2
  1
  3
  1
  2
  1
  2
  1
  2
  1
  2
  1
  2
  1
  9
  1
  8
            6
            1
            3
            1
           18
            1
            3
            4
       CONTENTS

RECORD TYPE
n I ti
TYPE OF CORRECTION
n I ii
TYPE OF BACKGROUND
n I n
BLANK
n I n
YEAR COMPUTED
N I it
MONTH COMPUTED
n I n
DAY COMPUTED
ii I n
CAS # OF INTERFERING ANALYTE
n i »
ANALYTE WAVELENGTH
         CORRECTION FACTOR

         EXPONENT
         ii i ii
         BLANK
         n i n
         RECORD SEQUENCE NUMBER
         RECORD CHECKSUM
 FORMAT

"35"

"ICP"/"BG
                                                               n 1
"ES"/"BD"/"BZ"



YY

MM

DD

CHARACTER

NUMERIC
(to t-.ro decir.^1

NUMERIC
"E"/BLANK
NUMERIC
                                  NUMERIC
                                  NUMERIC
  "ICP"  indicates interelement correction, while "BG"  indicates a backgrounc
  correction.
                                    H-42
                                                             ILM01.0

-------
              FORMAT  OF  THE SAMPLE HEADER DATA RECORD (TYPE 90V


COLUMNS   LENGTH           CONTENTS                    FORMAT

 1-2     2      RECORD  TYPE                        "90"
 3-3     1      " |"
 4-70    67      ANY  COMMENT                        CHARACTER
71-71     1      "|»
72-74     3      RECORD  SEQUENCE NUMBER            NUMERIC
75-78     4      RECORD  CHECKSUM                   NUMERIC
                                     H-43                    ILM01.0

-------
              FORMAT OF THE SAMPLE HEADER DATA RECORD  fTYPE  92)
COLUMNS    LENGTH          CONTENTS                     FORMAT

 1-2      2       RECORD TYPE                        "92"
 3-3      1       " | "
 4-12      9       COLOR BEFORE                       CHARACTER
13-13      1       "| "
14-22      9       COLOR AFTER                        CHARACTER
23-23      1       "| "
24-29      6       CLARITY BEFORE                     CHARACTER
30-30      1       "|"
31-36      6       CLARITY AFTER                      CHARACTER
37 - 37      1       "|"
38-43      6       TEXTURE                            CHARACTER
44-44      1       "| "
45-45      1       ARTIFACTS                           "YES"/BLANK
46-46      1       "|»
47 - 70    24       BLANK
71 - 71      1       " | "
72-74      3       RECORD SEQUENCE NUMBER             NUMERIC
75-78      4       RECORD CHECKSUM                    NUMERIC
                                     H-44                     ILM01.0

-------
9.   Example of  the Sequence of Record Types in a Production Run

10  Contains run header information.  Occurs once per run.

15  Contains additional run header information.  Occurs once per run.

20  Acts as a header for the following instrument parameters information.  Occurs
    once per run.  EPA SAMPLE NUMBER equals "SIDICF".  WAVELENGTH COUNTER equals
    the number of the type 32,  34, and 35 groups that follow.

       32  Contains integration time information for the wavelength on the type 34
          and 35 records that follow.  Occurs once for each wavelength used in the
          run.

       34  Contains the IDL and Linear range information for the first wavelength
          used in the run.

       35  Contains the background and interelenient correction information for the
          first wavelength used in the run.

       32  Contains integration time information for the wavelength on the type 34
          and 35 records that follow.  Occurs once for each v«»velength used in the
          run.
       34  Contains the IDL and Linear range information for second wavelength used
          in the run.
       25  Contains the 1 ackground and i-.^erelement correction information for
          second wavelength used in the run.

       32
       33

       35

          Continues as many times as the value of the WAVELENGTH COUNTER on the
          previous type 20  record.

20  Contains header information for sample and QC data.   Occurs as many times as
    there are entries on Form XIV for the run.

21  Contains additional information for analytical and instrument QC samples.
    '."ill always follow type 20  record.

22  Contains additional information for analytical samples.  Will usually follow
    type 21 record.   It is  not  required for instrument QC samples such as
    Instrument  Calibration  Standards (S), ICV,  ICB,  CCV,  CCB,  ICSA, ICSAB, CRI,
    and CRA.

28  Contains  additional information for analytical samples.  Will usually follow
    type 22 record.   It is  not  required for instrument QC samples such as
    Instrument  Calibration  Standards (S), ICV,  ICB,  CCV,  CCB,  ICSA, ICSAB, CRI,
    and CRA.

      30  Contains the sample level concentration,  true or added value and QC
          value for each analyte.   Occurs once for each analytical result for the
          EPA Sample  Number of  the previous type 20 record.

      31  Reports any instrumental data necessary to obtain the result reported on
          the previous type 30  record.   Will always follow type 30 record.  Occurs
          once  per type 30  record.
                                         H-45                        ILM01.0

-------
       33  Reports the average of instrumental data,  instrument related, qualifier,
          and the QC limits and qualifier for the QC VALUE reported on. previous
          type 30 record.  Will always follow type 31 record.   Occurs once per
          type 30 record.
       30  Values for the next analyte wavelength being measured.
       31  Values for the next analyte wavelength being measured.
       33  Values for the next analyte wavelength being measured.
       30
       31
       33
          Continues as many times  as the value of the WAVELENGTH  COUNTER on the
          previous type 20 record.
20  Next Sample Header record -  The following applies to the next sample data.
21
22
28
       30
       31
       33
       30
       31
       33    etc.
20
21
22
2£
      30
      31
      32
                  etc.
                                         H-46                         ILM01.0

-------