A540/R-93/051
             United States       Office of          Publication 9240.0-05A
             Environmental Protection   Solid Waste and       EPA540/R-93/051
             Agency         Emergency Response     PB93-983316
                        Washington, DC 20460     December 1992
             Superfund
    &EPA   Specifications and
             Guidance for
             Contaminant-Free
             Sample Containers

-------
                      TABLE OF CONTENTS
SECTION    TITLE                                               PAGE

 I.         INTRODUCTION  	 1

 II.         SAMPLE CONTAINER AND
           COMPONENT MATERIAL SPECIFICATIONS  	4

 III.         SAMPLE CONTAINER PREPARATION
           AND CLEANING PROCEDURES  	 14

 IV.        SAMPLE CONTAINER QUALITY ASSURANCE
           AND QUALITY CONTROL REQUIREMENTS 	 17
                              ill

-------
                                             SECTION I

                                          INTRODUCTION
        In August 1989, the Environmental Protection Agency's (EPA) Office of Emergency and Remedial
 Response (OERR) decentralized Superfund's Sample Container Repository program (OSWER Directive
 #9240.0-05).  In conjunction with the decentralization of Superfund's bottle program, OERR issued
 specifications and guidance for preparing contaminant-free sample containers to assist the Regions in
 obtaining appropriate sample containers from commercially available suppliers.

        The December  1992 version of "Specifications and Guidance for Contaminant-Free Sample
 Containers" revises the specifications and provides a single source of standardized specifications and
 guidance on appropriate cleaning procedures for preparing contaminant-free sample containers that meet
 all Contract Laboratory Program (CLP) detection/quantitation limits, including those for low
 concentration analyses.1  Although the specifications and guidance procedures contained in this document
 are based on CLP  low concentration requirements, they also are suitable for use in other analytical
 programs.

        Specifications and guidance for preparing contaminant-free sample containers are provided in  the
 sections that follow and are intended to describe one approach for obtaining cleaned, contaminant-free
 sample containers for use by groups performing sample collection activities under Superfund and other
 hazardous waste programs.  Although other cleaning procedures may be used, sample containers must
 meet the criteria specified in Section II.  In certain instances, the user of the sample containers may
 require exact adherence to the cleaning procedures and/or quality control analysis described  in this
 document.  In other instances, the user may require additional or different cleaning procedures and/or
 quality control analysis of the sample containers. The specific needs of the bottle user will determine  the
 requirements for the cleaning and quality control analysis of the sample containers as long as the
 minimum criteria are met.  It is the responsibility of the bottle user to define the sample container
 preparation, cleaning, and quality control requirements.

       The document has been  extensively reviewed and revised since the August 1989  iteration, and
 important enhancements have been incorporated, including:

       •    Removing references  to the color of the closures;

       •    Allowing the use of polypropylene closures as an alternative to phenolic closures;

       •    Referencing CLP Low Concentration Organics and Inorganics Statements  of Work for the
             analysis of calibration verification solutions and blanks;

       •    Including cleaning and quality control procedures for fluoride and nitrate/nitrite;

       •    Removing the hexane rinse from the cleaning procedure for container types A, E, F,  G,  H, J,
             and K (semivolatile organics,  pesticides, metals, cyanide, and fluoride in soils and water);
    1  Because this document does not address the procurement of contaminant-free sample containers, the
title was changed from "Specifications and Guidance for Obtaining Contaminant-Free Sample Containers"
to "Specifications and Guidance for Contaminant-Free Sample Containers."

-------
        •    Adding the recommendation that the bottle vendor establish and submit a Quality Assurance
             Plan (QAP);

        •    Changing the QA/QC documentation requirements so that copies of the raw data from the
             analyses of the QC containers are available upon request and not automatically sent to the
             bottle purchaser;

        •    Changing the permanent lot number assignment to a nine-digit number from an eight-digit
             number, where the extra digit represents the analysis parameter;

        •    Adding Chemical Abstract Services (CAS) registry number for the inorganic analytes in
             Table 1; and

        •    Recommending an annual demonstration of the bottle vendor's ability to meet detection
             limits and establish reproducibility of the cleaning  techniques.

        OERR and the EPA Regions decided to use the most stringent CLP requirements available to set
the specifications for obtaining contaminant-free sample containers.  As a result, the CLP Inorganics and
Organics Low Concentration Statement of Work (SOW) requirements were selected as the basis for these
specifications. Major factors in this decision included the desire to have a set of bottle cleaning
specifications that met or exceeded all analytical requirements and the related need  to avoid  potential
misuse of cleaned bottles (e.g., using a container cleaned by a multi-concentration procedure for a low
concentration sample).  OERR will reevaluate  this decision if the low concentration requirements are
deemed to be too stringent.

        Most environmental sampling and analytical applications offer numerous opportunities for sample
contamination. For this reason, contamination is a common source of error in environmental
measurements.  The sample container itself represents one such  source of sample contamination.  Hence,
it is vital that sample containers used within the Superfund program meet strict specifications established
to minimize contamination which could affect subsequent analytical determinations. Superfund sampling
and analysis activities require all component materials (caps, liners, septa, packaging materials, etc.)
provided by the bottle preparer to meet the criteria limits of the bottle specifications listed within Section
II.

        Section HI provides guidance on cleaning procedures for preparing contaminant-free sample
containers that meet the specifications contained in Section II. The procedures provided in this section
are intended to provide sample containers that meet all current CLP Low Concentration Inorganics and
Organics detection/quantitation levels.

        In selecting cleaning procedures for sample containers, it is important to consider all of the
parameters of interest.  Although a given cleaning procedure may be effective for one parameter or type of
analysis, it may be ineffective for another.  When multiple determinations are performed on  a single
sample or on a subsample from a single container, a cleaning procedure may actually be a source of
contamination for some analytes while minimizing contamination in others.  It should be the responsibility
of the bottle supplier to verify that the  cleaning procedures actually used satisfy the quality control
requirements set forth in Section IV.

-------
        Two aspects of quality assurance (i.e., quality control and quality assessment) must be applied to
sample containers as well as to  the analytical measurements.  Quality control includes the application of
good laboratory practices and standard operating procedures especially designed for the cleaning of sample
containers. The cleaning operation should be based on protocols especially designed for specific
contaminant problems.  Strict adherence to these cleaning protocols is imperative. Quality assessment of
the cleaning process depends largely on monitoring for adherence to the respective protocols.  Because of
their critical role in the quality assessment of the cleaning operation, protocols must be carefully designed
and followed.  Guidance is provided in Section IV on design and implementation of quality assurance and
quality control  protocols.

-------
                                          SECTION II

             SAMPLE CONTAINER AND COMPONENT MATERIAL SPECIFICATIONS
       This Section identifies sample containers commonly used in the Superfund program and provides
specifications for contaminant-free sample containers for each bottle type.

A.     CONTAINER MATERIAL

       A variety of factors affect the choice of containers and cap material.  These include resistance to
breakage, size, weight, interferences with analytes of interest, cost, and availability.

       Container types A through L (Figure 1, pages 6-7) are designated as  the type of sample containers
that have been used successfully in the past.  Kimax or Pyrex brand borosilicate glass is inert to most
materials and is recommended where glass containers are used (i.e., pesticides and other organics).
Conventional polyethylene is recommended  when plastic is acceptable because of its lower cost and lower
adsorption of metal ions.  The specific sampling situation will determine the use of plastic or glass.

       While the sample containers shown in Figure 1 are utilized primarily for Superfund sampling
activities, they also may be used for sampling activities under other programs, such as the Resource
Conservation and Recovery Act (RCRA).

B.     MAXIMUM CONTAMINANT LEVEL SPECIFICATIONS  FOR SAMPLE CONTAINERS

       The CLP, through a series of technical caucuses, has established inorganic Contract Required
Detection Limits (CRDL) and organic Contract Required  Quantitation Limits (CRQL) which represent
the minimum quantities needed to support the hazardous substance identification and monitoring
requirements necessary for remedial and other actions at hazardous waste sites.

       For inorganic sample containers,  the CRDLs listed in Table 1, page 8, are the specifications for
maximum trace metal contamination.  Concentration at or above these limits on any parameter should
preclude these containers from use in collecting inorganic samples.

       The CRQL specifications for organic sample containers are listed in  Table 2, pages 9-13.  When
the CRQL in Table 2 is multiplied by the appropriate factor listed below, the resulting value then
represents the maximum concentration allowed for particular sample  containers based on organic CLP
sample sizes for routine analyses.

          Container type                               Multiple of CRQL
               A                                         1.0
               B                                         0.5
               D                                        10.0
               E                                         8.0
               F                                         4.0
               G                                        2.0
               H                                        0.5
               J                                         0.5
               K                                        2.0

-------
       The philosophy used for determining the maximum permissible amount of contamination in a
sample container was to consider the number of aliquots of sample that are available in the containet and
assume that the contamination present would be uniformly distributed in all of the aliquots.  This
assumption, and the assumption that there should be no more than one-half the CRQL contributed by  the
container, resulted in the  establishment of contamination limits  by container type. For example, the
volume of container type  D is sufficient to allow 20 volatile determinations. Therefore, if 10 times the
CRQL of contaminant is  present in  the cleaned bottle, each aliquot tested will contain one-half of the
CRQL of contaminant due to the contribution from the bottle.

C.     GROSS CONTAMINATION

       Gross contamination is defined as greater than two hundred times the acceptable concentration
values in Tables 1 or 2 (multiplied by the appropriate factor), unless the cleaning procedure is successful
in reducing the  amount of contamination to within specifications.  If this is not achieved, the grossly
contaminated materials should be discarded and replaced to prevent cross contamination with other
batches of containers.  The bottle preparer should inspect  all materials to ensure  conformance with the
required specifications.

-------
                                             FIGURE I

                                       SAMPLE CONTAINER
                                         SPECIFICATIONS
Container
Type           Specifications
  A            Container:  80-oz amber glass, ring handle
               bottle/jug, 38-mjn neck finish.
               Closure: polypropylene or phenolic cap,
               38-430 size; 0.015-in Teflon liner.
               Total Weight:  2.45 Ibs.

  B            Container:  40-mL glass vial, 24-mm neck finish.
               Closure: polypropylene or phenolic, open-top,
               screw cap, 15-cm opening, 24-400 size.
               Septum: 24-mm disc of 0.005-in Teflon
               bonded to 0.120-in silicon for total thickness
               of0.125-in.
               Total Weight:  0.72 oz.

  C            Container:  1-L high-density polyethylene,
               cylinder-round  bottle, 28-mm neck finish.
               Closure: polyethylene cap, ribbed, 28-410 size;
               F217 polyethylene liner.
               Total Weight:  1.89 oz.

  D            Container:  120-mL wide mouth, glass vial,
               48-mm neck finish.
               Closure: polypropylene cap, 48-400 size;
               0.015-in Teflon liner.
               Total Weight:  4.41 oz.

  E            Container:  16-oz tall, wide mouth,
               straight-sided, flint glass jar,
               63-mm neck finish.
               Closure: polypropylene or phenolic cap,
               63-400 size; 0.015-in Teflon liner.
               Total Weight:  9.95 oz.

  F            Container:  8-oz short, wide mouth,
               straight-sided, flint glass jar,
               70-mm neck finish.
               Closure: polypropylene or phenolic cap,
               70-400 size; 0.015-in Teflon liner.
               Total Weight:  7.55 oz.

-------
                                             FIGURE 1

                                       SAMPLE CONTAINER
                                         SPECIFICATIONS
                                            (Continued)
Container
Type	Specifications
  G           Container:  4-oz tall, wide mouth,
               straight-sided, flint glass jar,
               48-mm neck finish.
               Closure: polypropylene or phenolic cap,
               48-400 size; 0.015-in Teflon liner.
               Total Weight:  4.70 oz.

  H           Container:  1-L amber, Boston round, glass
               bottle, 33-mm pour-out neck finish.
               Closure: polypropylene or phenolic cap,
               33-430 size; 0.015-in Teflon liner.
               Total Weight:  1.11 Ibs.

  J            Container:  32-oz tall,  wide mouth,
               straight-sided, flint glass jar,
               89-mm neck finish.
               Closure: polypropylene or phenolic cap,
               89-400 size; 0.015-in Teflon liner.
               Total Weight:  1.06 Ibs.

  K           Container:  4-L amber glass, ring handle
               bottle/jug, 38-mm neck finish.
               Closure: polypropylene or phenolic cap,
               38-430 size; 0.015-in Teflon liner.
               Total Weight:  2.88 Ibs.

  L           Container:  500-mL  high-density polyethylene,
               cylinder-round bottle, 28-mm neck finish.
               Closure: polypropylene cap, ribbed, 28-410 size;
               F217 polyethylene liner.
               Total Weight:  1.20 oz.

-------
                                    TABLE 1

                              INORGANIC ANALYTE
                                SPECIFICATIONS

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
Analyte
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Mercury
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
Cyanide
Fluoride
Nitrate/Nitrite
CAS Number
7429-90-5
7440-36-0
7440-38-2
7440-39-3
7440-41-7
7440-43-9
7440-70-2
7440-47-3
7440-48-4
7440-50-8
7439-89-6
7439-92-1
7439-95-4
7439-96-5
7439-97-6
7440-02-0
7440-09-7
7782-49-2
7440-22-4
7440-23-5
7440-28-0
7440-62-2
7440-66-6
57-12-5
16984-48-8
1-005
CRDL1 (/ig/L)
100
5
2
20
1
1
500
10
10
10
500
2
500
10
0.2
20
750
3
10
500
10
10
20
10
200
100
CRDLs are based on the CLP Inorganics Low Concentration SOW

-------
                                    TABLE 2

                               ORGANIC COMPOUND
                                 SPECIFICATIONS

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
Volatiles
Chloromethane
Bromomethane
Vinyl Chloride
Chloroethane
Methylene Chloride
Acetone
Carbon Disulfide
1,1-Dichloroethene
1,1-Dichloroethane
cis- 1 ,2-Dichloroethene
trans-l,2-Dichloroethene
Chloroform
1,2-Dichloroethane
2-Butanone
Bromochloromethane
1,1,1 -Trichloroethane
Carbon Tetrachloride
Bromodichloromethane
1 ,2-Dichloropropane
cis- 1 ,3-Dichloropropene
Trichloroethene
Dibromochloromethane
1 , 1 ,2-Trichloroethane
Benzene
trans-l,3-Dichloropropene
Bromoform
4-Methyl-2-pentanone
2-Hexanone
Tetrachloroethene
1 , 1 ,2,2-Tetrachloroethane
CAS Number
74-87-3
74-83-9
75-01-4
75-00-3
75-09-2
67-64-1
75-15-0
75-35-4
75-34-3
156-59-4
156-60-5
67-66-3
107-06-2
78-93-3
74-97-5
71-55-6
56-23-5
75-27-4
78-87-5
10061-01-5
79-01-6
124-48-1
79-00-5
71-43-2
10061-02-6
75-25-2
108-10-1
591-78-6
127-18-4
79-34-5
CRQL1 (Mg/L)
1
1
1
1
2
5
1
1
1
1
1
1
1
5
1
1
1
1
1
1
I
1
1
1
1
1
5
5
1
1
1 CRQLs are based on the CLP Organics Low Concentration SOW

-------
                                   TABLE 2 (cont.)

                                ORGANIC COMPOUND
                                   SPECIFICATIONS

31.
32.
33.
34.
35.
36.
37.
38.
39.
40.
Volatiles
1,2-Dibromoethane
Toluene
Chlorobenzene
Ethylbenzene
Styrene
Xylenes (total)
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1 ,2-Dichlorobenzene
1 ,2-Dibromo-3-chloropropane
CAS Number
106-93-4
108-88-3
108-90-7
100-41-4
100-42-5
1330-20-7
541-73-1
106-46-7
95-50-1
96-12-8
CRQL1 (Mg/L)
1
1
1
1
1
1
1
1
1
1
1 CRQLs are based on the CLP Organics Low Concentration SOW
                                          10

-------
                                   TABLE 2 (cont.)

                                ORGANIC COMPOUND
                                  SPECIFICATIONS

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
Semivolatiles
Phenol
bis-(2-Chlorethyl)ether
2-Chlorophenol
2-Methylphenol
2,2'-oxybis-(l-Chloropropane)
4-Methylphenol
N-Nitroso-di-n-dipropylamine
Hexachloroethane
Nitrobenzene
Isophorone
2-Nitrophenol
2,4-Dimethylphenol
bis-(2-Chloroethoxy)methane
2,4-Dichlorophenol
1 ,2,4-Trichlorobenzene
Naphthalene
4-Chloroaniline
Hexachlorobutadiene
4-Chloro-3-methylphenol
2-Methylnaphthalene
Hexachlorocyclopentadiene
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
Dimethylphthalate
Acenaphthylene
2,6-Dinitrotoluene
3-Nitroaniline
Acenaphthene
CAS Number
108-95-2
111-44-4
95-57-8
95-48-7
108-60-1
106-44-5
621-64-7
67-72-1
98-95-3
78-59-1
88-75-5
105-67-9
111-91-1
120-83-2
120-82-1
91-20-3
106-47-8
87-68-3
59-50-7
91-57-6
77-47-4
88-06-2
95-95-4
91-58-7
88-74-4
131-11-3
208-96-8
606-20-2
99-09-2
83-32-9
CRQL1 (Mg/L)
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
20
5
20
5
5
5
20
5
1 CRQLs are based on the CLP Organics Low Concentration SOW
                                         11

-------
                                   TABLE 2 (cont.)

                               ORGANIC COMPOUND
                                  SPECIFICATIONS

31.
32.
33.
34.
35.
36.
37.
38.
39.
40.
41.
42.
43.
44.
45.
46.
47.
48
49.
50.
51.
52.
53.
54.
55.
56.
57.
58.
59.
60.
Semivolatiles
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2,4-Dinitrotoluene
Diethylphthalate
4-Chlorophenyl-phenylether
Fluorene
4-Nitroaniline
4,6-Dinitro-2-methylphenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenylether
Hexachlorobenzene
Pentachlorophenol
Phenanthrene
Anthracene
Di-n-butylphthalate
Fluoranthene
Pyrene
Butylbenzylphthalate
3,3'-Dichlorobenzidine
Benz[a]anthracene
Chyrsene
bis-(2-Ethylhexyl)phthalate
Di-n-octylphthalate
Benzo[b]fluoranthene
Benzo[kjfluoranthene
\
Benzo[a]pyrene
Indeno(l,2,3-cd)pyrene
Dibenz[a,h]anthracene
Benzo[g,h,i]perylene
CAS Number
51-28-5
100-02-7
132-64-9
121-14-2
84-66-2
7005-72-3
86-73-7
100-01-6
534-52-1
86-30-6
101-55-3
118-74-1
87-86-5
85-01-8
120-12-7
84-74-2
206-44-0
129-00-0
85-68-7
91-94-1
56-55-3
218-01-9
117-81-7
117-84-0
205-99-2
207-08-9
50-32-8
193-39-5
53-70-3
191-24-2
CRQL1 (Mg/L)
20
20
5
5
5
5
5
20
20
5
5
5
20
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
1 CRQLs are based on the CLP Organics Low Concentration SOW
                                         12

-------
                                   TABLE 2 (cont.)

                                ORGANIC COMPOUND
                                  SPECIFICATIONS

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
Pesticides/PCBs
alpha-BHC
beta-BHC
delta-BHC
gamma-BHC (Lindane)
Heptachlor
Aldrin
Heptachlor epoxide
Endosulfan I
Dieldrin
4,4'-DDE
Endrin
Endosulfan II
4,4'-DDD
Endosulfan sulfate
4,4'-DDT
Methoxychlor
Endrin ketone
Endrin aldehyde
alpha-Chlordane
gamma-Chlordane
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
CAS Number
319-84-6
319-85-7
319-86-8
58-89-9
76-44-8
309-00-2
1024-57-3
959-98-8
60-57-1
72-55-9
72-20-8
33213-65-9
72-54-8
1031-07-8
50-29-3
72-43-5
53494-70-5
7421-36-3
5103-71-9
5103-74-2
8001-35-2
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
CRQL1 (Mg/L)
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.02
0.02
0.02
0.02
0.02
0.02
0.02
0.10
0.02
0.02
0.01
0.01
1.0
0.20
0.20
0.40
0.20
0.20
0.20
0.20
1 CRQLs are based on the CLP Organics Low Concentration SOW
                                         13

-------
                                           SECTION III

              SAMPLE CONTAINER PREPARATION AND CLEANING PROCEDURES


       This Section is provided as guidance for the preparation of sample containers that meet the
contaminant-free specifications contained in Section II.  There are various procedures for cleaning sample
containers depending upon the analyses to be performed on the sample. The following cleaning
procedures are modeled after those specified for the Superfund Sample Container Repository program.
Other suitable cleaning procedures exist and may be used as long as the sample containers meet the
criteria established in Section II.  In some instances, the specific needs of the bottle user may dictate exact
adherence to the sample container preparation and cleaning procedures that  follow; while in other
instances, modifications may be required.  It is the responsibility of the bottle user to define the sample
container preparation, cleaning, and quality  control requirements.

A.     Cleaning Procedure for Container Types:  A, E, F, G, H, J, and K

1.      Sample Type: Semivolatile Organics, Pesticides, Metals, Cyanide, and Fluoride in Soils and Water.

       a.   Wash glass bottles, Teflon liners, and caps with  hot tap water using laboratory grade
            nonphosphate detergent.

       b.   Rinse three times with copious amounts of tap water to remove detergent.

       c.   Rinse with 1:1  nitric acid  (reagent  grade HNO3, diluted with ASTM Type I deionized water).

       d.   Rinse three times with ASTM Type I organic free water.

       e.   Oven dry bottles, liners, and caps at  105-125°C for one  hour.

       f.    Allow bottles, liners, and caps to cool to room temperature in  an enclosed contaminant-free
            environment.

       g.   Rinse bottles with pesticide grade methylene chloride (or other suitable solvents specified by
            the bottle user) using 20 mL for 1/2-gallon containers; 10 mL for 32-oz and 16-oz containers;
            and 5 mL for 8-oz and 4-oz containers.

       h.   Oven dry bottles, liners, and caps at  105-125°C for one hour.

       i.    Allow bottles, liners, and  caps to cool to room temperature in  an enclosed contaminant-free
            environment.

       j.    Place liners in  lids and cap containers.

       k.   Label each container with the lot number and pack in a case.

       1.    Label exterior  of each case with the  lot number.

       m.  Store in a contaminant-free area.
                                                 14

-------
2.      Sample Type:  Nitrate/Nitrite in Soils and Water.

        a.    Substitute reagent grade sulfuric acid (H2SO4) for nitric acid in step A.I.e.

        b.    Follow all other steps in the cleaning procedure described in part A. I above.

B.      Cleaning Procedure for Container Types:  B, D

1.      Sample Type:  Purgeable (Volatile) Organics in Soils and Water.

        a.    Wash glass vials, Teflon-backed septa. Teflon liners, and caps in hot water using laboratory
             grade nonphosphate detergent.

        b.    Rinse three times with copious amounts of tap water to remove detergent.

        c.    Rinse three times with ASTM Type I organic-free water.

        d.    Oven dry vials, caps, septa, and liners at 105-125°C for one hour.

        e.    Allow vials, caps, septa, and liners to  cool to room temperature in an enclosed contaminant-
             free environment.

        f.    Seal 40-mL vials with septa (Teflon side down) and cap.

        g.    Place liners in lids and cap 120-mL vials.

        h.    Label each  vial with the lot number and pack in a case.

        i.    Label exterior of each case with the lot number.

        j.    Store in a contaminant-free area.

C.      Cleaning Procedure for Container Types: C, L

1.       Sample Type: Metals, Cyanide, and Fluoride in Soils and Water.

        a.    Wash polyethylene bottles and caps in hot tap water using laboratory-grade nonphosphate
             detergent.

        b.    Rinse three times with  copious amounts of tap water to remove detergent.

        c.    Rinse with  1:1 nitric acid (reagent grade HNO3,  diluted with ASTM Type I deionized water).

        d.    Rinse three times with  ASTM Type I deionized water.

        e.    Invert and air dry in a contaminant-free environment.

        f.    Cap bottles.

        g.    Label each container with the lot number and pack in a case.
                                                 15

-------
        h.    Label exterior of each case with the lot number.




        i.    Store in a contaminant-free area.




2.       Sample Type:  Nitrate/Nitrite in Soils and Water.




        a.    Substitute reagent grade sulfuric acid (H2SO4) for nitric acid in step C.l.c.




        b.    Follow all other steps in the cleaning procedure described in part C.1 above.
                                                  16

-------
                                           SECTION IV

     SAMPLE CONTAINER QUALITY ASSURANCE AND QUALITY CONTROL REQUIREMENTS


A.      Quality Assurance

        The objectives of this Section are to: (1) present procedures for evaluating quality assurance (QA)
information to ensure that specifications identified in Section II have been met; and (2) discuss techniques
for the quality control (QC) analysis of sample containers to be used in conjunction with the cleaning
procedures contained in Section III.

        The bottle vendor should establish a Quality Assurance Plan (QAP) with the objective of
providing sound analytical chemical measurements, production procedures, and tracking systems. The
QAP should incorporate procedures for the inspection of incoming raw materials; preparation, cleaning,
and labeling of container lots; quality control analyses of cleaned container lots; document control,
including all documentation required for analysis, packing, shipping, and tracking of container lots; any
necessary corrective actions; and any quality assessment measures implemented by management to ensure
acceptable performance. The QAP should be available and provided to the bottle purchaser upon  request.

        Major QA/QC activities should include the inspection of all incoming materials, QC analysis of
cleaned lots of containers, and monitoring of the container storage area. Complete documentation of all
QC inspection results  (acknowledging acceptance or rejection) should be kept as part of the permanent
bottle preparation files.  QA/QC records (e.g., preparation/QC logs, analytical data, data tapes, storage log)
also should be stored in a central location within the facility.

        Documentation  indicating that the container lot has passed all QA/QC requirements should be
provided by the bottle vendor to the bottle purchaser with each container lot. Documentation should
include a signed and dated cover statement affirming that all QA/QC criteria were met.  Copies of raw
data from applicable analyses of the QC containers, laboratory standards, check samples, and blanks should
be available and provided upon request.  Original documentation should be retained for at least 10 years.
Minimum documentation that should be available, if applicable, for each lot of containers includes:

        •   A statement that "Sample container lot	meets or exceeds all QA/QC criteria
            established in 'Specifications and Guidance for Contaminant-Free Sample Containers;'"

        •   Reconstructed Ion Chromatographs (RICs) from volatile and semivolatile organics
            determinations, including calibration verification standards, check samples, and blanks;

        •   GC chromatographs from pesticides determinations, including calibration verification
            standards,  check samples, and blanks;

        »   ICP, hydride-ICP, or  ICP-MS instrument readouts from  metals determinations, including
            calibration verification standards, check samples, and blanks;

        •   AA raw  data sheets and instrument readouts from metals determinations, including
            calibration verification standards, check samples, and blanks; and

        •   Cyanide, fluoride, and nitrate/nitrite raw data sheets and instrument readouts from these
            determinations, including calibration verification standards, check samples, and blanks.
                                                 17

-------
        Prior to the first shipment of containers, and at least annually thereafter, the bottle vendor should
demonstrate its ability to meet the CRDLs and CRQLs, and establish the reproducibility of the cleaning
techniques for each bottle type. The ability to meet the CRDLs and CRQLs is accomplished through the
determination of instrument detection limits (IDLs). The bottle vendor should use the procedures in the
current  CLP Low Concentration Inorganics and Organics SOWs to determine IDLs.  IDLs should be
below the CRDLs or CRQLs.  To establish the reproducibility for each bottle type, the bottle vendor
should randomly pick seven containers from a cleaned lot and analyze as described in the Quality Control
Analysis part of this Section.  Parameter concentrations should  be at or below the CRDL or CRQL  for
each bottle type.  Documentation from these analyses should be available and provided upon request.

1.      Incoming Materials Inspection:

        A representative item from  each case of containers should be checked for conformance with
specifications provided in Section II. Any deviation should be considered unacceptable.  A log of
incoming shipments should be maintained to identify material type, purchase order number, and delivery
date.  The date of incoming inspection and acceptance or  rejection of the material should also be  recorded
on this  log.

2.      Quality Control Inspection of Cleaned Lots of Containers:

        Following container cleaning and labeling, containers should be randomly selected from each
container lot to be used for QC purposes. The two categories of  QC containers should be as follows:

        a.    Analysis QC Containers:

        One percent of the total number of containers in each  lot should be designated as the analysis QC
        container(s).  For lots of less than 100 containers, one  container should be designated as the
        analysis QC container.  The sample container preparer  should analyze the analysis QC
        container(s)  to check for contamination prior to releasing the associated container lot for
        shipment. The QC analyses procedures specified  in the Quality Control Analysis part of this
        Section for determining the presence of semivolatile and volatile organics, pesticides, metals,
        cyanide, fluoride, and nitrate/nitrite should be utilized.

        For each analysis QC container(s), an appropriate QC  number should be assigned that cross-
        references the  QC container to the related lot of containers. For example, the QC number  could
        be a seven-digit number sequentially assigned to each lot  that has undergone QC analysis. Under
        this numbering scheme, the first alphabetical character  would be the container type letter from
        Figure 1, the next four digits would be assigned sequentially in numerical order starting with
        "0001" for the first lot to undergo QC analyses, the sixth character would indicate the number of
        QC container for the lot, (e.g., "1" for the first QC container in the lot, "2" for the second, etc.)
        and the last character would be either a "C" to indicate clearance or an "R" to indicate rejection.

        If the representative analysis QC container(s) passes QC  inspection, the related lot of containers
        should be released, and the appropriate QC number should be entered in the preparation/QC log
        to indicate clearance of the lot for shipment.

        If the analysis  QC container(s) are found to be contaminated per the specified QC analysis
        procedures, the appropriate QC rejection number should be assigned and entered in the
        preparation/QC log.  Any container labels should be removed and the entire lot returned for
        reprocessing under a new lot number.  Excessive  QC rejection for a particular container type
        should be noted for future reference.
                                                 18

-------
        A laboratory standard, check sample, and a blank should be run with each QC analysis.  A
        calibration verification standard should be analyzed once every  12 hours.  All QC analysis results
        should be kept in chronological order by QC report number in  a central QC file.  The QC
        numbers  assigned should be documented in the preparation/QC log, indicating acceptance or
        rejection  and date of analysis.

        A container lot should not be released for shipment prior to QC analysis and clearance. Once the
        containers have passed QC inspection, the containers  should be stored in a contaminant-free area
        until packaging and shipment.

        b.    Storage QC Containers:

        One QC  container per lot should be designated as the storage QC container. The storage QC
        container should be separated from the lot after cleaning and labeling and should be stored in a
        designated contaminant-free area for one year.  The date the container is placed in the storage
        area should be recorded in the storage QC container log.

        If contamination of the particular container lot comes into question at any time following
        shipment, the storage QC container should be removed  from the storage area and analyzed  using
        the  QC analysis procedures for that container type (see  Quality Control Analysis, this Section).
        Upon removal, containers should be logged out of the storage area.

        The designated storage area should be monitored continuously for volatile contaminants in  the
        following manner.  A precleaned, 40-mL vial  that has  passed a QC inspection should be filled with
        ASTM Type I organic-free water and be placed in the storage area. This vial should be changed at
        one-week intervals.  The removed vial should be subjected to analysis for volatile organics as
        described in the Quality Control Analysis part of this  Section.  Any peaks  indicate contamination.
        Identify contaminants, if present, and include the results in a report to all  clients who purchased
        bottles from the affected lot(s).

B.      Quality Control Analysis

        The types of QC analyses correlate with the types of containers being   analyzed and their future
use in sample collection. The QC analyses are intended  for the  determination of:

        •    Semivolatile organics and pesticides;

        •    Volatile organics;

        •    Metals;

        •    Cyanide;

        •    Fluoride; and

        •    Nitrate/Nitrite.

        QC  analyses should be performed according to the container type and related sample type and
utilize the specific method(s) described below.
                                                 19

-------
1.      Determination of Semivolatile Organics and Pesticides:

       Container Types: A, E, F, G, H, J, and K

       a.   Sample Preparation:

                Add 60 mL of pesticide-grade methylene chloride to the container and shake for two
                minutes.

                Transfer the solvent to a Kuderna-Danish (KD) apparatus equipped with a three-ball
                Snyder column.  Concentrate to less than 10 mL on a steam bath. Split the solvent  into
                two 5 mL fractions  for semivolatile and pesticide determinations.

                Add 50 mL of pesticide-grade hexane (for pesticide determinations  only) to the KD
                apparatus by slowly pouring down through the Snyder column.  Concentrate to less than
                10 mL to effect solvent replacement of hexane for  methylene chloride.

                Concentrate the solvent to I mL using a micro-Snyder column.

                Prepare a solvent blank by adding 60 mL of the rinse solvent used in step "g" of the
                cleaning procedure for container types A, E, F, G, H, J, and K (Section III page 14)
                directly to a KD apparatus, and proceed  as above.

       b.   Semivolatile Organics Sample Analysis:

                Instrument calibration should be performed as described in the  most recent CLP Low
                Concentration Organics SOW with the following exceptions:

                (1)   If problems are encountered meeting the %RSD criteria on the initial calibration
                     for semivolatiles, the high concentration point should be deleted and a four-point
                     calibration used.

                (2)   The low concentration standard should be used for the continuing calibration
                     standard for semivolatile analyses.

                (3)   The percent difference window should be widened  to ± 30 percent for all
                     compounds.

                Inject 1 (iL of solvent into a gas chromatograph/mass spectrometer (GC/MS).

                Calibration verification standards should be analyzed as  described in the most recent
                CLP Low  Concentration Organics SOW.

                Blanks should be run as described in the most recent CLP Low Concentration Organics
                SOW.

                If compounds other than those listed in Table 2 are found in the container blank that
                are not in the solvent blank at a peak height or peak area greater than  20 percent of the
                nearest internal standard, the containers should be rejected (See Section  II, Table 2  for
                compound specifications).

                Identify and quantitate any contaminant(s) that cause rejection of a  container lot.

                                                20

-------
                 A standard mixture of the nine semivolatile organic compounds listed in Table 3 (nage
                 26) with concentrations in the 5-20 ppb range should be analyzed to ensure that
                 sensitivities are achieved that will meet contract required quantitation limits.  This
                 standard should be prepared from a different source from the calibration standards.

        c.    Pesticides Sample Analysis:

                 Instrument calibration should be performed as described in the most recent CLP Low
                 Concentration Organics SOW.

                 Inject 1 IJ.L of solvent into a gas chromatograph (GC) equipped with an electron capture
                 detector (ECD).

                 Calibration verification standards should be analyzed  as described in the most recent
                 CLP Low Concentration Organics SOW.

                 Blanks should be run as described in the most recent CLP Low Concentration Organics
                 SOW.

                 If compound peaks other than those listed in Table 2 are  at a peak height or peak area
                 greater than 5 percent of the peak height or peak area of tetra chloro-m-xylene, the
                 containers should be  rejected (See Section II, Table 2).

                 Identify and quantitate any contaminant(s) that cause rejection of a container lot.

                 A standard mixture of the seven pesticide compounds listed  in Table 3 (page 26) with
                 concentrations in the 0.01  to 1 ppb range should be analyzed to ensure that sensitivities
                 are achieved that will meet contract required quantitation limits.  This standard should
                 be prepared from a different source from the calibration standards.

2.       Determination of Volatile Organics:

        Container Types:  B and D

        a.    Sample Preparation:

                 Fill the container with ASTM Type I organic-free water.

                 Cap the container and let stand  for 48 hours.

        b.    Sample Analysis:

                 Instrument calibration should be performed as described in the most recent CLP Low
                 Concentration Organics SOW with the following exceptions:

                 (1)   If problems are  encountered meeting the %RSD criteria on the initial calibration
                      for volatiles, the high concentration point should be deleted and a four-point
                      calibration used.

                 (2)   The low concentration  standard should be used  for the continuing calibration
                      standard for volatile analyses.

                 (3)   The percent difference  window should be widened to ± 30 percent.

                                                 21

-------
                Calibration verification standards should be analyzed as described in the most recent
                CLP Low Concentration Organics SOW.

                Blanks should be run as described in the most recent CLP Low Concentration Organics
                SOW. The blank should consist of an aliquot of the ASTM Type I water used in the
                sample preparation.

                If compounds other than those listed in Table 2 are found in the container blank that
                are not in the solvent blank at a peak height or peak area greater than 20 percent of the
                nearest internal standard, the containers should be rejected (See Section  II, Table 2 for
                compound specifications).

                Identify and quantitate any contaminant(s) that cause rejection of a container lot.

                A standard mixture of the five volatile organic compounds listed in Table 3 (page 26)
                with concentrations in the 1-5 ppb range should be analyzed to ensure that sensitivities
                are achieved  that will meet contract required  quantitation limits.  This standard should
                be prepared from a different source from the  calibration standards.

3.       Determination of Metals:

        Container Types:  A, C, E, F, G, H, J, K and  L

        a.    Sample Preparation:

                Add 100 mL of ASTM Type I deionized water to the container,  and acidify with 1.0 mL
                of reagent-grade HNO3.  Cap and shake for three to five minutes.

                Cap the container and let stand for  48 hours.

                Treat the sample as a dissolved metals sample.  Analyze the undigested water using the
                most recent CLP Low Concentration Inorganics SOW.

        b.    Sample Analysis:

                Instruments used for the analysis of the samples should meet the contract required
                detection limits in Table 1.

                The ASTM Type I deionized water should be analyzed before use on  the bottles that are
                designated for analysis to ensure that contaminated water is not used for rinsing the
                bottles.

                Calibration verification standards should be analyzed as described in the most recent
                CLP Low Concentration Inorganics SOW.

                Blanks should be analyzed as described in the most recent CLP  Low Concentration
                Inorganics SOW.  A calibration blank is a solution made up exactly like the sample
                preparation solution. The calibration blank should be less than the values contained in
                Table 1.

                A set of standards in the expected working range should  be analyzed with each analytical
                run.  The acid matrix of the standards, blank, and quality control samples should match
                that of the samples.

                                                22

-------
                Concentrations at or above the detection limit for each parameter (listed in Table I)
                should be cause for rejection of the lot of containers.  NOTE: The sodium detection
                limit for container types A, E, F,  G, H, J, and K is 5000 ^g/L unless the containers will
                be used for low concentration analyses, then the detection limit is 500 ng/L.

4.       Determination of Cyanide:

        Container Types: A, C, E,  F, G, H, J, K and L

        a.   Sample Preparation:

                Place 250 mL of ASTM Type I deionized water in the container.  Add  1.25 mL of 6N
                NaOH (for container types F and G use 100 mL of ASTM Type I deionized water and
                0.5 mL of 6N NaOH).  Cap the container and shake vigorously for two minutes.

        b.   Sample Analysis:

                Analyze an aliquot as described in the most recent CLP Low Concentration Inorganics
                SOW.

                The detection limit should be 10  /ig/L or lower.

                Calibration verification standards should be analyzed as described in the most recent
                CLP Low Concentration Inorganics SOW.

                Blanks should be run as described in the  most recent CLP Low Concentration Inorganics
                SOW.  The calibration  blank should consist of an aliquot of the ASTM Type I water
                used above.

                A set of standards in the expected working range, a check sample, and blank should be
                prepared exactly as the  sample was prepared.

                The detection of 10 /jg/L cyanide (or greater) should be cause for rejection of the lot of
                containers. NOTE: Contamination could be due to the container, the cap, or the
                NaOH.

5.       Determination of Fluoride:

        Container Types: A, C,  E, F, G, H, J, K and L

        a.   Sample Preparation:

                Place 250 mL of ASTM Type I deionized water in  the container (for container types F
                and G use 100 mL of ASTM Type I deionized water). Cap the container and shake
                vigorously for two  minutes.

        b.   Sample Analysis:

                Analyze an aliquot as described in the most recent CLP  Low Concentration Inorganics
                SOW.

                The detection limit should be 200 ^g/L or lower.
                                                23

-------
                Calibration verification standards should be analyzed as described in the most recent
                CLP Low Concentration Inorganics SOW.

                Blanks should be run as described in the most recent CLP Low Concentration Inorganics
                SOW. The calibration blank should consist of an aliquot of the ASTM Type I water
                used above.

                A set of standards in the expected working range, a check sample, and blank should be
                prepared exactly as the sample was prepared.

                The detection of 200 pg/L (or greater) of fluoride should be cause for rejection of the lot
                of containers. NOTE: Contamination could be due to the container or the cap.

6.       Determination  of Nitrate/Nitrite:

        Container Types: A, C, E, F, G, H, J, K and L

        a.    Sample Preparation:

                Place 250 mL of ASTM Type I deionized water in the container  (for container types F
                and G use 100 mL of ASTM Type I deionizec. water). Cap the container and shake
                vigorously for two minutes.

        b.    Sample Analysis:

                Analyze an aliquot as described in the most recent CLP Low Concentration Inorganics
                SOW.

                The detection limit should be 100 /zg/L or lower.

                Calibration verification standards should be analyzed as described in the most recent
                CLP Low Concentration Inorganics SOW.

                Blanks should be run as described in the most recent CLP Low Concentration Inorganics
                SOW. The calibration blank should consist of an aliquot of the ASTM Type I water
                used above.

                A set of standards in the expected working range, a quality  control sample, and blank
                should be prepared exactly as the  sample was prepared.

                The detection of 100 ^g/L (or greater) of nitrate/nitrite should be cause for rejection of
                the lot of containers.  NOTE: Contamination could be due  to the container or the cap.

C.       Preparation and Labeling

        Sampling for environmental specimens requires  that sample containers be  transported to field sites
prior to sample collection.  As  a result, considerable time may elapse between the  receipt of sample
containers and collection of the samples.  Because of the large number of samples  taken at any one site,
accounting for all sample containers can become extremely difficult. The following guidance on the
identification and tracking of sample containers is based on procedures that have been used successfully in
the CLP bottle program.
                                                24

-------
1.       Each shipment should be inspected to verify that the requested number of cleaned and prepared
        sample containers have been supplied and meet the requirements specified in Section II (Tables 1
        and 2). If any shipment fails to meet the required  specifications, it should be discarded and
        replaced with a supply of sample containers that meet the required criteria.

2.       The sample containers should be removed and prepared in accordance with the methods
        designated below.

3.       A permanent nine-digit lot number should  be assigned to each lot of sample containers for
        identification and tracking purposes throughout the life of the containers.  Figure 2 provides an
        example of a lot number sequence.

                                            FIGURE 2

                                    LOT NUMBER SEQUENCE
                            95th day
                               of    —i
                            the year
—Analysis Parameter
                                                              _Repository
                                                              Code
                                  2   095  01
                Container  	|
                 Type A
                          Year 1992
                                             _Belongs to the  1st Lot
                                              washed that day
       a.    The first digit represents the container type in Section II (Figure 1).

       b.    The second digit represents the last digit of the calendar year.

       c.    The next three digits represents the day of the year on which the sample containers were
            washed.

       d.    The sixth and seventh digits represent the daily  lot number.

       e.    The eighth digit represents the analysis parameter where:

            A = Semivolatile organics, pesticides, metals, cyanide, and fluoride;
            B = Metals,  cyanide, and fluoride;
            V = Volatile organics;
            S  = Semivolatile organics and/or pesticides;
            M = Metals;
            C = Cyanide;
            F = Fluoride; and
            N = Nitrate/nitrite.

       f.    The final digit represents the identification of the person who prepared the lot.
                                                25

-------
4.      The lot number for each container should be entered, along with the date of washing, type of
       container, and number of containers per lot, into the preparation/QC log book.

5.      Lot numbers printed with solvent resistant ink on a nonremovable label should remain with the
       corresponding containers throughout the cleaning procedure.

6.      After sample container cleaning and drying, the label should be affixed to the containers in a
       permanent manner.

7.      At least one face should be clearly marked, excluding the top and bottom faces, of each case of
       sample containers with the assigned lot numbers.
                                            TABLE 3

          STANDARD MIXTURES OF ORGANIC COMPOUNDS TO VERIFY SENSITIVITY
       Volatiles
Semivolatiles
Pesticides
       Methylene Chloride
       Acetone
       2-Butanone
       Trichloroethene
       Toluene
Nitrobenzene
4-Chloroaniline
2,6-Dinitrotoluene
Diethylphthalate
4-Bromophenyl-phenylether
Hexachlorobenzene
Pentachlorophenol
Di-n-butylphthalate
bis(2-Ethylhexyl)phthalate
Gamma-BHC
Heptachlor
Aldrin
Dieldrin
Endrin
4,4'-DDT
Aroclor 1260
                                               26
                                   *U.S. G.P.O.:1993-341-835:81056

-------