PA747/R-95-001
                   United States
                   Environmental Protection
                   Agency
Pollution Prevention
and Toxics
(7404)
EPA747-R-95-001
March 1995
                   Residential Sampling for Lead:
                   Protocols for Dust and Soil Sampling
                   Final Report
                                                 Recycled/Recyclable
                                                 Printed with Soy/Canola Ink on paper that
                                                 contains at least 50% recycled fiber

-------
       Residential Sampling for Lead:
Protocols for Dust and Soil Sampling

                          Final Report
    For U.S. Environmental Protection Agency
     Office of Pollution Prevention and Toxics
        Technical Programs Branch, MC-7407
                 Northeast Mall, Room G012
                        401 M Street, S.W.
                   Washington, D.C. 20460

                     Attn: Darlene Watford
                 Work Assignment 4-10(02)
                      MRI Project No. 9803

                           March  29, 1995

-------
                              Disclaimer
This document has been reviewed and approved for publication by the Office of
Prevention and Toxics, U.S. Environmental Protection Agency.  The use of trade
names or commercial products does not constitute Agency endorsement or
recommendations for use.

-------
                      Table of Contents
A.    Introduction  	   1
            1.0    Overview  	   1
            2.0    Purpose	   1
            3.0    Related Documents	   1
            4.0    Number and Location of Samples	   2
            5.0    Sample Collection  	   2
            6.0    Laboratory Analysis	   3

B.    Protocol for Collection of Soil Samples for Lead Determination  ....   6
            1.0    Introduction	   6
            2.0    Equipment and Supplies  	   6
            3.0    Sampling Procedure  	   8
            4.0    Quality Control 	  11
            5.0    Glossary	  13

C.    Protocol for Collection of Dust Samples for Lead
      Determination Using Wipe Sampling  	  15
            1.0    Introduction	  15
            2.0    Equipment and Supplies  	  15
            3.0    Sampling Procedure  	  16
            4.0    Quality Control 	  20
            5.0    Glossary	  23

D.    Protocol for Collection of Dust Samples for Lead
      Determination Using Vacuum Sampling 	  25
            1.0    Introduction	  25
            2.0    Equipment and Supplies  	  25
            3.0    Sampling Procedure  	  27
            4.0    Quality Control 	  33
            5.0    Glossary	  36

E.    References	  38

-------
A.    Introduction

1.0   Overview

This document provides guidance for the collection of soil and settled dust samples for
subsequent determination of lead.  Collection of soil samples is performed using either
coring or spooning methods. Collection of settled dust samples are performed using
wipe or vacuum methods.  Advantages and disadvantages of each method are
discussed.  Protocols presented in this document are capable of producing samples
for lead determination results in jig per gram of soil, (ig per gram of dust, and u,g
per ft2 of dust.
2.0   Purpose

The purpose of this document is to provide detailed sampling procedures to maintain
uniformity in data collection of lead in soil and settled dust samples. Analytical results
obtained from samples collected using these sampling procedures will eventually be
used for comparison to numerical health-based standards developed by the U.S.
Environmental Protection Agency (EPA) for discerning unsafe lead levels, according to
Title X, Section 403: Identification of Dangerous Levels of Lead.

Today, a variety of approaches to sample collection for lead determinations are being
utilized. Substantial resources are applied to the generation of lead data.  The use of
standardized protocols should improve the value of lead data gathered by a wide
variety of organizations and individuals.
3.0   Related Documents

Currently, EPA is developing health-based standards for lead in soil and house dust
under the Toxic Substances Control Act (TSCA), Title X, Section 403. Furthermore,
EPA will soon publish standards on the conduct of residential lead risk assessments
for the purposes of training risk assessors (TSCA, Title IV, Section 402).  Both
standards are interwoven and extremely important in addressing public health issues
regarding  lead.

For guidance to identify, control, or abate lead hazards in housing, consult the U.S.
Department of Housing and Urban Development document, Guidelines for the
Evaluation and Reduction of Lead-Based Paint Hazards in Housing (HUD. 1994).

To obtain  more  information on lead, call the National Lead Information Center
Clearinghouse at (800)  424-LEAD. In the Washington, D.C., area, call (202)
833-1071.

-------
Protocols presented in this document are equivalent to ASTM methods1"2 for the
collection of soil and dust.
4.0   Number and Location of Samples

A sampling plan is needed to determine where and how many samples are to be
collected at any given sampling site (or housing area).  Sampling plans typically
include a variety of sampling locations, including collection of soil samples from near
building locations (drip-line locations) and away from building locations, collection of
dust samples from painted friction surfaces (such as window areas), and other high
traffic areas.  A complete sampling plan is developed under  a specific lead
determination objective and should be  made prior to any sample collection. Specific
sampling plan guidance is provided in the EPA document Residential Sampling for
Lead: A Process for Risk Assessment. 1994.
5.0   Sample Collection

The amount of lead in soil and settled dust samples can be expressed as a lead
loading (weight of lead per area sampled, typically u-g/ft2),  or lead concentration
(weight of lead per weight of sample, typically u.g/g).  Soil  samples are generally
reported in concentration results, and dust samples can be reported using both
measures.  Currently, more data are available relating blood lead levels to dust
loading than to lead dust concentration.  However, since there are limits in the amount
of dust and soil material that can be ingested (or inhaled)  by an individual, a lead
value that incorporates the amount of dust, such as lead concentration, may provide
important information for accurate evaluation of lead hazards.  Unfortunately, technical
difficulties exist for collection of dust with the intent to report lead  results in
concentration units. These difficulties are discussed in this document.  In summary,
the collection method should be selected according to the desired lead measurement
unit because all collection methods do not produce results both in lead loading and
concentration terms.

The collection protocol for soil  samples is presented in Section B.  The protocol, which
provides for either scoop or coring type sampling options,  focuses on collection  of the
top 0.5 inches of soil surface.  It is assumed that the top 0.5 inches represents the
portion most likely to cause an immediate hazard to occupants.

Two collection protocols are presented for lead in settled dust: wipe dust and vacuum
dust collection.  These protocols are presented in Sections C and D. Both protocols
are intended for collection of settled dust as opposed to airborne paniculate. Settled
dust is generally referred to as dust material in the range of 5 to 500 ujn in  diameter.
However, these collection protocols are not  necessarily limited to  collection  of material
in this range. Wipe dust collection is limited to reporting lead results in loading terms

-------
(jig/ft2).  Vacuum dust collection can generate both lead loading and lead
concentration results.  However, vacuum dust collection is generally more difficult and
costly to perform than wipe dust collection.
6.0   Laboratory Analysis

Samples collected for lead determination are generally analyzed using atomic
spectrometry methods.  These instrumental methods require samples to be converted
from a solid to a liquid form prior to lead measurement. This conversion process is
commonly referred to as sample preparation. Sample preparation generally includes
several initial handling steps, followed by a digestion process, that solubilize the lead
contained in the sample.  The accuracy and  precision of the lead determination is
dependent on both the sample preparation and  instrumental analysis activities.  An
estimate of accuracy and precision can be obtained by submitting Quality Control (QC)
samples to the laboratory together with field  samples. Preparation of appropriate QC
samples  is discussed in the sample collection protocols presented in this document.
Although this document is not intended as a guide on laboratory analysis procedures,
a brief discussion of sample preparation and instrumental analysis activities, as they
relate to the collection of soil and settled dust sample, is presented below.
6.1    Sample Preparation of Soil Samples

A number of laboratory sample preparation methods can be used for the preparation
of soil samples.  Hot plate digestions, such as SW846 method 30503 or ASTM
ES S6-944, utilize nitric acid (sometimes with hydrochloric acid) and hydrogen peroxide
for oxidation of sample components and solubilization of lead.  Microwave digestions,
such as  30513, utilize nitric acid for oxidation of sample components and solubilization
of lead. Sample size limits exist for both digestion methodologies.  Samples collected
in the field and submitted to the laboratory are generally much larger than the
digestion methods are capable of accommodating.  Therefore, a subsample of the
original soil sample must be used for the lead determination.  This requires the use  of
a homogenization process prior to digestion to generate a representative subsample.
The water content of collected soils can be widely variable.  Therefore, to provide
consistency for lead data  comparisons under variable soil and weather conditions, soil
samples must be reported on  a dry weight basis.  This  requires inclusion of a drying
process into sample preparation method used for  soil analysis.

In summary, any sample preparation method used for lead determination in soil
samples must include  a homogenization and drying process prior to subsampling for
digestion and solubilization of  the lead.

-------
6.2   Sample Preparation of Wipe Dust Samples

Wipe dust samples can be prepared in a manner similar to soil samples (i.e., digestion
using acids and hydrogen peroxide). However, microwave digestion methods are not
generally applicable because the large bulk of the wipe material used to collect the
samples interferes with the digestion process. Modifications to standard hot plate
digestion methods, such as SW846 method 30503, are generally required to assure
that the wipe samples are completely covered with the digestion reagents (acids and
hydrogen peroxide) during the digestion process. The ASTM dust wipe digestion,
ES 36-944, is a modification of SW846 method 30503.  Unlike soil samples, wipe dust
samples cannot be subsampled; therefore, no homogenization procedures are
required for preparation of these samples.

In summary, any sample preparation method used for lead determination in wipe dust
samples must include sufficient reagent volumes to completely digest both the wipe
itself and the collected dust.
6.3   Sample Preparation of Vacuum Dust Samples

Vacuum dust samples also can be prepared in a manner similar to soil samples (i.e.,
hot plate or microwave digestion using acids or hydrogen peroxide, or both).
However, if these samples are intended for reporting lead results in concentration
units (u.g/g) then the total sample weight must be determined, and the sample
preparation procedure must include a specific sample weighing  procedure. This
reporting option requires a determination of the total collected sample weight.  This
determination is performed by inclusion of gravimetric procedures before field
sampling (called prefield gravimetrics) and after field sampling (called postfield
gravimetrics).  Both prefield and postfield gravimetrics must be performed under
identical humidity and temperature conditions to help minimize errors caused by the
hydroscopic nature of the sample media (filters or filter cassettes).  Similar to wipe
dust samples, vacuum dust samples cannot be subsampled; therefore, no
homogenization procedures are required for preparation of these samples.

In summary, any sample preparation method used for lead determination in vacuum
dust samples that are intended for reporting lead results in concentration units (u.g/g)
must include prefield and postfield weighing procedures, quantitative transfer
procedures,  and digestion procedures.
6.4   Instrumental Measurement

Lead measurements in digested soil and dust samples can be performed using a
variety of instrumental techniques. Atomic spectrometry methods commonly used
include Flame Atomic Absorption Spectrometry (FAAS), Graphite Furnace Atomic

-------
Absorption Spectrometry (GFAAS), and Inductively Coupled Plasma Atomic Emission
Spectrometry (ICPAES). A variety of methods covering the use of these techniques,
such as SW846 methods 74203, 74213 and 60103, or ASTM E 1613-945, can be used
for lead measurements. In general,  FAAS and ICPAES both have sufficient  detection
capability for lead determinations in all soil samples and most dust samples.
However, for clean environments, GFAAS detection,  which has approximately a
10-fold improvement in detection capability over FAAS and ICPAES, may be more
appropriate for some dust samples.  Lead measurement data must incorporate sample
preparation variables, such as sample weights and digestion volumes, prior to
reporting lead results.

-------
B.  Protocol for Collection of Soil Samples for Lead Determination

1.0 Introduction

This protocol provides for the collection of soil samples using either scooping or coring
methods.  The protocol  is applicable for collection of soil samples for lead
determination.


2.0 Equipment and Supplies

2.1 Scoop Sampling Equipment

    2.1.1    Plastic centrifuge tube. 50-mL with screw-on  cap.  Used for scoop
            sampling or soil collection containers.

    2.1.2    Sample  collection container, resealable plastic bags (1  quart or 1  gallon)
            or sealable rigid-walled container with  50-mL minimum volume.

    2.1.3    Spoon, plastic or stainless steel. Used for scoop sampling.

    2.1.4    Steel or plastic measuring tape or ruler, divisions to at  least 1/s inch.


2.2 Core Sampling Equipment

    2.2.1    Coring probe. 0.5 inch minimum diameter, lead-free. The probe must be
            capable of being forced into hard ground to a depth of at least 2 inches
            without being damaged and have a mechanism to remove the core from
            the probe  to permit discarding all but the top 0.5 inch of the soil core
            (see subsection 2.2.2). A number of devices can be utilized as a coring
            probe.  Examples include: plastic or steel pipe, small tree sapling
            planters, and  a professional stainless  steel coring probe equipped with
            plastic liners,  cross T-bar, and hammer.

    2.2.2    Coring plungers, one with and one without a stop, sized to fit coring
            probe, lead-free.  Removal of the soil  core is generally performed using a
            pair of plungers machined to fit the inside diameter of the coring device.
            One plunger is equipped with a stop that limits extension of the plunger
            to within 0.5 inch from the far end of the coring  probe.  It is used to
            remove  all except the top 0.5 inch of the soil core from the coring probe.
            The other  plunger (without a stop) is used to remove the remaining
            0.5 inch of the soil core from the coring probe.

-------
    2.2.3    Sample collection container, resealable plastic bags (1  quart or 1 gallon)
            or sealable rigid walled container with 50-mL minimum  volume. If plastic
            bags are used, samples should be double bagged to protect against
            breakage and potential sample loss.

    2.2.4    Spoon, plastic or stainless steel. Used for scoop sampling.

    2.2.5    Steel or plastic measuring tape or ruler, divisions to at  least Va inch.


2.3 General Supplies

    2.3.1    Field notebooks, bound with individually numbered pages, see
            subsection 4.1.

    2.3.2    Indelible ink marker, black or blue.

    2.3.3    Ink pens, black or blue.

    2.3.4    Packaging tape, used for  sealing shipping containers.

    2.3.5    Plastic baas, trash bags with ties.

    2.3.6    Plastic gloves, powderless.  Gloves with powder should not be used to
            avoid potential contamination of samples from powder  material.

    2.3.7    Preprinted field forms,  preprinted with sufficient entry lines to address
            documentation needs presented in subsection 4.1

    2.3.8    Shipping containers, cardboard or plastic for interim storage and
            shipment of sample collection containers.


2.4.     Cleaning Supplies

    2.4.1    Water, drinking water.  Drinking water is used to assist in cleaning
            sampling equipment for soil sample collection.  High purity  water is  not
            required for cleaning of sampling equipment because action levels for
            lead in soils are relatively high with  respect to lead levels in drinking
            water.

    2.4.2    Wipe.  Disposable towelette moistened with a wetting agent.  Used  for
            cleaning sampling equipment.   Wipe brands or sources should contain
            insignificant background lead levels.  Laboratory analysis of replicate
            blank wipes should be used to determine background lead  levels prior to

-------
            use in the field.  It is recommended to avoid brands of wipes that contain
            aloe because wipes containing aloe have been found to contain higher
            background lead levels.  Background lead levels less than 10 u.g per
            wipe are considered insignificant for most soil sampling activities.
3.0 Sampling Procedure

Two types of collection procedures are described in this section: scoop sampling and
core sampling.  Either procedure can be used for the collection of soil samples for
lead determinations.  Advantages and disadvantages of each are presented at the
beginning of each procedure.
3.1 Scoop Sampling Procedures

Two procedures are provided for collection of soils at a given sampling location using
a scooping methods.  Scooping procedures are effective for collection from semisoft,
sticky, and loose,  sandy soils.  Scooping procedures are not recommended for hard or
frozen soils. Scooping procedures are less effective than coring methods for
collection of multiple samples having uniform surface area sampled and consistent
sampling depths.  The scooping methods described here may result in collection bias
toward increased  amounts of surface soil as opposed to subsurface soil caused by the
curvature of the scooping tools. Coring methods are generally free from this collection
bias.

    3.1.1    Scoop Sampling Using a  Plastic Centrifuge Tube

        3.1.1.1 Label a new plastic 50-mL centrifuge tube for use as a sample
        collection container (See subsections 2.1.2 and 5.5).

        3.1.1.2 Pull on a pair of clean, powderless, plastic gloves.  Gloves are used
        to protect the workers' hands and the integrity of the samples (to aid in
        avoiding cross-contamination between samples).

        3.1.1.3 Using a measuring tape and a spare plastic 50-mL centrifuge tube,
        determine the proper scooping depth of the tube needed to collect
        approximately the top 0.5 inch soil. For example, if the plastic centrifuge tube
        is about 1 inch in diameter, then the proper scooping depth is to insert the
        tube into  the soil until the soil surface is about  even with the center of the
        tube.

        3.1.1.4 Remove the cap of the plastic centrifuge tube and insert the open
        end of the tube into  the soil to the desired depth as  determined in
        step 3.1.1.3.  Collect the soil into the tube by pushing  or pulling the tube
                                       8

-------
    through the soil surface while maintaining the scooping depth of the tube
    (0.5 inch) in the soil.  Move the tube a distance of 6-12 inches across the soil
    surface to complete collection of the soil into the tube.  The movement of the
    tube across the sample location will result in a composite type soil sample.

    3.1.1.5 Remove the tube from the  ground, and wipe off any excess soil
    clinging to the outside of the tube and cap threads with a gloved finger.
    Replace the cap.  Label the plastic  centrifuge tube with sufficient information
    to uniquely identify the sample. Discard any gloves used during  sample
    collection in a trash bag.

3.1.2    Scoop Sampling Using a Spoon

    3.1.2.1 Label a new resealable plastic bag for use as a sample collection
    container (See subsection  5.5).

    3.1.2.2 Pull on a pair  of clean, powderless, plastic gloves.  Gloves are  used
    to protect the workers' hands and the integrity of the samples (to aid in
    avoiding cross-contamination between samples).

    3.1.2.3 Using a measuring tape and a clean spoon,  dig a small test hole
    adjacent to the sampling location to the depth of 0.5  inch.  Use this hole as a
    visual aid during soil collection to help limit collection to a depth of 0.5 inch.
    Clean the spoon using a wipe until  soil is no longer visible on the spoon.

    3.1.2.4 Scoop the  soil with the spoon down to the depth indicated by the test
    hole and place the sample in a sample collection  container.  Continue to
    collect soil until a circular hole of approximately 2 inch (0.5 inch deep) has
    been created.

    3.1.2.4 Collect soil from two more  locations within a 1 foot diameter circle
    around the first sample location, using the same procedure described above
    (subsections 3.1.2.2 through 3.1.2.4). Composite these scoop samples into
    the same sample collection container and seal the container in a manner that
    will minimize the air contained in the container.  Discard any gloves  used
    during collection  in a trash bag after all three scoop samples have been
    collected and composited.

    3.1.2.5 Pull on a pair  of clean, powderless, plastic gloves.  Clean the spoon
    using wipes and water until soil is no  longer visible on the spoon. Discard
    any wipes and gloves  used during cleaning in a trash bag.  An alternative
    approach to cleaning is to use disposable spoons.

-------
3.2 Core Sampling Procedures

The collection of soils using a coring method at a given sample location is provided in
this subsection.  Coring methods are effective for collection of soils from dense, hard,
or sticky soils. Coring methods are not recommended for loose, sandy soils.  Coring
methods generally produce samples with more uniform surface areas and consistent
sampling depths than scooping methods.

    3.2.1  Label a new resealable plastic bag for use as a sample collection container
    (See section 5.5).

    3.2.2  Pull on a pair of clean, powderless, plastic gloves. Gloves are used to
    protect the workers' hands and the integrity of the samples (to aid in avoiding
    cross-contamination between samples).

    3.2.3     If needed, clean the coring probe and coring plungers using wipes or
    water.  The sampling equipment is considered clean if no soil or other debris is
    visible on any of the surfaces.  Check the stop on the coring plunger, equipped
    with a stop, to ensure that the plunger tip stops at a distance of 0.5 inch from the
    end of the coring probe.  Adjust the stop if needed.

    3.2.4     Place a directional arrow on the outside of the coring probe with the
    arrow head pointed toward the ground.  This arrow identifies the orientation of the
    soil core with respect to the surface of the ground. The arrow is used to avoid
    inadvertent loss of the top of the soil core when the plunger is used to remove
    and collect the soil sample. If the coring probe is a professional stainless steel
    coring tool equipped with plastic liners, place the arrow on the outside of the
    plastic liner and orient the liner in the probe so that the arrow head is pointed
    toward the ground.

    3.2.5     Grip the coring tool firmly between two hands and, using a slight twisting
    motion,  drive the tool into the soil surface at the designated sampling location to a
    depth of at  least 2  inches.  The directional arrow (from section 3.2.4) must be
    pointing down.  For extremely hard soils (i.e., hard packed  or frozen), a hammer
    or other similar device may be needed to drive the tool into the ground.  If
    conditions do not allow for full penetration to a minimum  of 2 inches, make every
    effort to penetrate to a depth of at least 0.5 inches.  If penetration is less than
    0.5 inches,  the documentation generated for the sample  should indicate the
    approximate depth achieved.

    3.2.6     Twist and snap the coring tool to one side and carefully remove the tool
    from the ground while retaining the soil  core in the tool.

    3.2.7     Insert a clean plunger, equipped with stop,  into the top end of the coring
    probe or liner. (The bottom end is indicated by the arrow head drawn on the tool.
                                       10

-------
    The top end is the opposite opening.) Push out all but 0.5 inch of the soil from
    the probe with the plunger.  Using a gloved finger, wipe off the excess soil
    protruding from the probe. Allow the soil pushed out of the probe to fall on the
    ground near but not on the sampling location.

    3.2.8    Using a clean plunger (without stop), push the remaining 0.5 inch section
    of the sample core into a sample collection container.

    3.2.9    Collect two more soil cores within a 1 ft diameter circle around the first
    sampling location, using  the same procedure described above (subsections 3.2.2
    through 3.2.8). Composite these cores into the same  sample collection container
    and seal the container in a manner that minimizes the air contained in  the
    container. Discard any gloves used during collection in a trash bag after all three
    core samples have been collected and composited.

    3.2.10 Pull on a pair of  clean, powderless, plastic gloves. Clean the coring
    probe, coring plungers, and plastic inserts (if used) using wipes and water until
    soil is no longer visible on the equipment. Discard any wipes and gloves used
    during cleaning in a trash bag.
4.0 Quality Control

Adherence to quality control (QC) procedures is an important part of field sample
collection.  QC procedures, including documentation requirements, field QC samples,
reference material check samples, and contamination avoidance are presented in this
section.
4.1 Documentation

All field data related to sample collection must be documented.  A field notebook or
sample log form can be used to record field collection data.  It is recommended that
both types of documentation records (field notebooks and preprinted sample log
forms) be utilized to assure collection of all relevant field data. Field data entries on
documentation records must adhere to the following requirements:

    4.1.1    General Documentation  Requirements:

       •   All entries must be made using ink.
       •   Each  page (notebook or form)  must include the name of the person
            making the entries and the date of entries found on the page.
       •   Any entry errors must be corrected by using only a single line through
            the incorrect entry (no scratch  outs) accompanied by the initials of the
            person making the correction and the date of correction.
                                       11

-------
       •    An initial page that correlates initials to a specific name must be
            generated and maintained with field data records to trace any initials
            used in  notebooks and on data forms.

    4.1.2    Specific Sampling Site Documentation Requirements:

       •    General sampling site description.
       •    Project or client name, address, and city/state location.
       •    Information as to what specific collection protocol was used.
       •    Information as to the use of interim storage and sample shipment
            mechanisms.

    4.1.3    Documentation Required for Each Sample Collected:

       •    An individual and unique sample identifier and date of collection.  This
            must be recorded on the sample container in addition to the field data
            records (notebook or form).
       •    Name of person collecting the sample and specific sampling  location
            data from which the sample was removed.
4.2 QC Samples

    4.2.1    Blank Samples.  Normally, blank samples should be periodically collected
            (designated) throughout the sampling day at each sampling site.  Field
            blank samples are used to identify any potential systematic lead
            contamination present in the sampling media and handling of samples
            during field collection and laboratory analysis activities. However,
            because soil samples are not collected on a sampling media such as a
            wipe of filter, there is no practical method for collection of a blank
            sample.

            Although sampling equipment rinses can be used to collect potential field
            contamination related information, difficulties exist in laboratory
            processing of these "field blanks" with soil samples. Sampling equipment
            rinse-type field blanks cannot be carried through the homogenization/
            drying steps that are commonly applied to soil samples. In addition, lead
            results from equipment rinses are reporting in weight-volume units (i.e.,
            u,g/mL) and cannot be directly compared or related to the reported soil
            lead weight-weight results (i.e., |ig/g). Therefore, no field blanks are
            recommended for soil sampling. Contamination effects should be
            minimized through adherence to the procedures specified in this protocol.
                                       12

-------
    4.2.2    Blind Reference Material Samples.  Reference materials should be
            periodically submitted to the laboratory for analysis as a check on
            adherence to proper laboratory sample preparation and instrumental
            analysis methods.  Prepare a blind reference material by placing a
            portion (1-2 grams) of a reference material into a labeled sample
            collection container.  It is recommended that the frequency of these QC
            samples be at least 1 per 20 field samples.  Reference materials from
            MIST6, such as SRMs 2709, 2711, and 2704, are readily available and
            can be used for preparing blind reference materials.  Other sources of
            materials with known lead levels, such as soil materials from the ELPAT7
            program, may also be used as blind reference materials.
4.3 Contamination Avoidance

The following work practices should be followed to prevent cross-contamination of
samples:

        •    Avoid tracking soil from one location to another by:
                identifying and clearly marking all sampling locations upon arrival at
                the sampling site, and
                instructing field team members to avoid walking through or over any
                of the marked sampling location areas.
        •    Use a new pair of powderless gloves at each sampling location.
        •    Inspect all sampling equipment for cleanliness prior to collection of each
            sample. Always  clean suspect equipment if in doubt.
        •    Do not open sample containers until needed to collect each sample.
        •    When using bulk packed wipes, at each sampling location, discard the
            first two wipes pulled from the wipe container.
5.0 Glossary

    5.1  Digestion. Sample preparation process that solubilizes lead present in the
        sample.  The digestion process produces an acidified, aqueous solution called
        the digestate.  A lead determination is made on the digestate during an
        instrumental measurement process.

    5.2  Field Data. Any information collected at the sampling site.

    5.3  Field Sample.  Physical material taken from the sampling site that is targeted
        for lead determination.

    5.4  Reference Material. Material of known composition containing a known
        amount of lead.  These materials have typically been subjected to a large
                                      13

-------
    number of lead determinations to develop a lead result known to a high
    degree of confidence.

5.5 Sample Collection Container. Container for holding and transporting the
    samples from the field to the laboratory.  The internal volume of the container
    must be sufficient to hold the entire collected sample.

5.6 Sampling Location. Specified area within a sampling site that is subjected to
    sample collection.  Multiple sampling locations are commonly designated for a
    single sampling site.  An example would be at the bottom of a specific slide in
    a specific playground area.

5.7 Sampling Site. Local geographical area that contains the sampling locations.
    A sampling site is generally limited  to an area that can be easily covered on
    foot.  An example would be John Smith's house at 3102 Nowhere Avenue,
    Detroit, Ml.

5.8 Sample Preparation. Process used to ready a sample received from the field
    for lead determination using instrumental measurement methods.  The
    process is dependent on the sample type and can include a large number of
    steps such as homogenization, drying, splitting, weighing, digestion, dilution to
    a final known volume, and filtering.
                                  14

-------
C.  Protocol for Collection of Dust Samples for Lead Determination
    Using Wipe Sampling

1.0 Introduction

This protocol provides for the collection of settled dust samples from hard, relatively
smooth, nonporous surfaces using wipe methods.  The protocol is not applicable for
the collection of settled dust samples from highly textured surfaces, such as brickwork
and rough concrete, and soft fibrous surfaces, such as upholstery and carpeting. The
protocol is capable of producing samples for lead determination results in loading
terms (u,g/ft2).
2.0 Equipment and Supplies

2.1 Sampling Equipment

    2.1.1    Disposable shoe covers (optional), see subsection 4.3.

    2.1.2    Masking tape, used for holding down sampling templates and marking
            sampling locations.

    2.1.3    Sample collection  container, sealable rigid-walled container with 50-mL
            minimum volume.   Use of a resealable plastic bags for holding and
            transporting the settled dust wipe sample is not recommended due to the
            potential losses of settled dust within the plastic bag during laboratory
            handling.  Quantitative removal and processing of the settled dust wipe
            sample by the laboratory is significantly improved through the use of
            sealable rigid walled containers.

    2.1.4    Sampling template. 1 ft2 inside area reusable aluminum or plastic,  or
            disposable cardboard or plastic template. A variety of shapes  are
            recommended for  use in variable field situations such as square,
            rectangular, square "U" shaped,  rectangular "U" shaped, and "L."  All
            templates  must  have accurately known inside dimensions.  Templates
            should be thin (less than Vs inch) and capable of lying flat on a flat
            surface.

    2.1.5    Steel or plastic measuring tape or ruler, divisions to Vie inch.

    2.1.6    Wipe, disposable towelette moistened with a wetting agent.  Wipe
            brands or sources should contain insignificant background lead levels.
            Laboratory analysis on  replicate blank wipes should be used to
            determine background lead  levels prior to use in the field. Background
            lead levels less  than 10 jig per wipe are considered insignificant for most


                                      15

-------
            dust sampling activities.  It is recommended to avoid brands of wipes
            that contain aloe because wipes containing aloe have been found to
            contain higher background lead levels. Increased laboratory sample
            preparation difficulties have also been noted for wipes containing lanolin.
            Wipe brands or sources selected for use should be of adequate width
            and thickness to perform the collection procedure.  A thin wipe which is
            approximately 6 inches by 6 inches is recommended. Use of large,
            multiply or extra thick wipes can cause problems with laboratory analysis.
            Use  of wipes with smaller dimensions may not be capable of holding
            settled dust contained within the sampling area.
2.2 General Supplies

    2.2.1    Field notebooks, bound with individually numbered pages, see
            subsection 4.1.

    2.2.2    Indelible ink marker, black or blue.

    2.2.3    Ink pens, black or blue.

    2.2.4    Packaging tape, used for sealing shipping containers.

    2.2.5    Plastic bags, trash bags with ties.

    2.2.6    Plastic gloves, powderless. Gloves with powder should not be used to
            avoid potential contamination of samples from powder material.

    2.2.7    Preprinted field forms, preprinted with sufficient entry lines to address
            documentation needs presented in subsection  4.1

    2.2.8    Shipping containers, cardboard or plastic for interim storage and
            shipment of sample collection containers.


3.0 Sampling Procedure

Two sampling  procedures are presented. One is to accommodate collection of a
settled dust sample in an unrestricted area such as a floor (Template Assisted
Sampling Procedure). The other is to accommodate collection of a settled dust sample
in a restricted  area such as a window channel (Confined Area Sampling  Procedure).
The Confined Area Sampling Procedure should only be used when the Template
Assisted Sampling Procedure can  not be used due to sampling  location constraints.
The Confined Area Sampling Procedure assumes the operator can be orientated to a
collection position where the sampling location's width is greater than its  depth.  It also
                                       16

-------
         that the depth is no larger than the dimensions of a wipe.  If this is not true,
then the Template Assisted Sampling Procedure should be used.
3.1 Template Assisted Sampling Procedure

Following is a summary of this procedure:

    1.   Select a sampling location.
    2.   Mark the sampling location using a template.
    3.   Perform first wiping: Side-to-side, fold the wipe.
    4.   Perform second wiping: Top-to-bottom, fold the wipe.
    5.   Perform third wiping: Clean-up the corners, fold the wipe, and store the
        sample.

The detailed procedure is as follows:

    3.1.1     Pull on a pair of clean, powderless,  plastic gloves.

    3.1.2    Carefully place a clean template on  the surface in manner that minimizes
             disruption of settled dust at the sampling location.  Either tape or place a
             heavy object on the outside edge of the template to prevent it from
             moving during sample collection.  An alternative to using a template is to
             mark an outline of the sampling location using masking tape as
             described in subsection 3.2.2.

    3.1.3    Discard any gloves used to mark the area in a trash bag and  pull on a
             new pair of clean, powderless, plastic gloves.

    3.1.4    At the beginning of a sampling  period (or if a new bulk-packed container
             of wipes is opened), remove a minimum of the top 2 wipes from the
             container and wipe off gloved fingers with each wipe as they are
             removed.  Use the next wipe from the container to collect the sample.

    3.1.5    First Wiping.  Side-to-Side:  Hold one edge of the wipe between the
            thumb and forefinger, draping the wipe over the fingers of a gloved hand.
             Hold fingers together, hand flat, and wipe the selected surface area,
             starting at either corner  furthest away from the operator (referred to as a
            far corner), using a slow side to side (left-to-right or right-to-left)
             sweeping motion. During wiping, apply pressure to the finger tips.

            At the end of the first pass from one side to the other, turn the leading
             edge of the wipe (the portion of the  wipe touching the surface) 180
             degrees, pulling the wipe path slightly closer to the operator and make a
                                       17

-------
            second side-to-side pass in the reverse direction, slightly overlapping the
            first pass. (The 180 degree turn is used to assure that the wiping motion
            is always performed in the same direction on the wipe to maximize dust
            pickup.)  Continue to cover the sampling area within the template, using
            the slightly overlapping side-to-side passes with the 180 degree turns at
            each edge until the close corner  of the template is reached. Carefully lift
            the leading dust line into the wipe using a slight rolling motion of the
            hand to capture the dust inside the wipe.  Fold the wipe in  half with the
            sample side folded inside the fold.

    3.1.6    Second Wiping. Top-to-Bottom:  Using a clean side of the wipe,  perform
            a second wiping over the sampling area within the template starting from
            a far corner in the same manner used for the first wiping, except  use a
            top-to-bottom sweeping of the surface.  When the close corner of the
            template is reached, carefully lift  the leading dust line into the wipe using
            a slight rolling motion of the hand to capture the dust inside the wipe.
            Fold the  wipe  in half (again) with the sample from this second wiping
            folded inside the fold.

    3.1.7    Third  Wiping. Clean Corners:  Using a clean side of the wipe, perform a
            third wiping around the perimeter of the sampling area within the
            template to pick up any dust remaining  in the corners. Start from  one
            edge  of the template and use the same wiping technique as described
            above.  When the perimeter has  been wiped and the starting location
            reached, carefully lift the leading dust line into the wipe using a slight
            rolling motion  of the hand to capture the dust inside the wipe. Fold the
            wipe in half one more time with the sample from this third wiping folded
            inside the fold.

    3.1.8    Insert the folded wipe into a sample collection container. Using a tape
            measure, verify the internal dimensions of the sampling template used to
            collect the sample and label the sample collection container with
            sufficient information to uniquely  identify the sample and the dimensions
            of the selected dust sampling area (with units such as inches). Discard
            any gloves in the trash bag. If the template is a reusable type, clean the
            template with several clean wipes.
3.2 Confined Area Sampling Procedure

Following is a summary of this procedure:

    1.   Select a sampling location.
    2.   Mark the sampling location using masking tape.
    3.   Perform first wiping: One direction, Side-to-side, fold the wipe.
                                       18

-------
    4.   Perform second wiping: One direction (reverse), Side-to-side, fold the wipe.
    5.   Perform third wiping: Clean-up the corners, fold the wipe, and store the
        sample.

The detailed procedure is as follows:

    3.2.1    Pull on a pair of clean, powderless, plastic gloves.

    3.2.2    Mark an outline of the sampling location using masking tape.  Care
            should be taken to minimize any disruption of dust at the sampling
            location. For areas that are dirty or contain high dust levels, new tape
            may have to be applied more than once to get adhesion to the surface.
            Discard any soiled tape in a trash bag.

    3.2.3    Discard any gloves used to mark the  area in a trash bag and pull on a
            new pair of clean, powderless,  plastic gloves.

    3.2.4    At the beginning of a sampling  period (or if a new bulk-packed container
            of wipes is opened), remove a  minimum of the top 2 wipes from the
            container and wipe off gloved fingers  with each wipe as they are
            removed. Use the next wipe from the container to collect the sample.

    3.2.5    First Wiping.  One Direction. Side-to-Side:  Hold one edge of the wipe
            between the thumb and forefinger, draping the wipe over the fingers  of a
            gloved hand.  Hold fingers together, hand flat, and wipe the selected
            surface area, starting at either corner furthest away from the operator
            (referred to as  a far corner), using a slow side to  side (left-to-right or
            right-to-left) sweeping motion.   During wiping, apply pressure to the
            finger tips.  At the end of the first pass from one side to the other,
            carefully lift the leading dust line into the wipe using a slight rolling
            motion of the hand to capture the dust inside the  wipe.  Fold the wipe in
            half with the sample side folded inside the fold.

    3.2.6    Second Wiping. One Direction. Side-to-Side:  Using a clean side of the
            wipe, repeat step 3.2.5 using a wiping motion in the reverse direction.

    3.2.7    Third Wiping. Clean Corners:  Using a clean side of the wipe, perform a
            third wiping around the perimeter of the sampling area to collect any dust
            remaining in the corners.  Start from the middle of one edge of the area
            and use the same wiping technique as described above. When the
            perimeter has been wiped and  the starting location  reached, carefully lift
            the leading dust line into the wipe using a slight rolling motion of the
            hand to capture the dust inside the wipe.  Fold the wipe in  half one more
            time with the sample from this third wiping folded inside the fold.
                                       19

-------
    3.1.8    Insert the folded wipe into a sample collection container. Using a tape
            measure, measure the dimensions of the sampled area and label the
            sample collection container with sufficient information to uniquely identify
            the sample and the dimensions of the selected sampling area (with units
            such as inches). Discard any gloves in the trash bag.
4.0 Quality Control

Adherence to quality control (QC) procedures is an important part of field sample
collection.  QC procedures, including documentation requirements, field QC samples,
reference material check samples, and contamination avoidance are presented in this
section.
4.1  Documentation

All field data related to sample collection must be documented. A field notebook or
sample log form can be used to record field collection data. It is recommended to
utilize both types of documentation records (field notebooks and preprinted sample log
forms) for assuring collection of all relevant field data. Field data entries on
documentation records must adhere to the following requirements:

    4.1.1    General Documentation Requirements:

        •   All entries must be made using ink.
        •   Each page (notebook or form) must include the name of the person
            making the entries and the date of entries found on the page.
        •   Any entry errors must be corrected by using only a single line through
            the incorrect entry (no scratch outs), and marked with the initials of the
            person making the correction and the date of correction.
        •   An initial page that correlates initials to a specific name must be
            generated and maintained with field data records to trace any initials
            used in notebooks and on data forms.

    4.1.2   Specific Sampling Site Documentation Requirements:

        •   General sampling site description.
        •   Project or client name, address, and city/state location.
        •   Information as to what specific collection protocol was used.
        •   Information as to the use of interim storage and sample shipment
            mechanisms.
                                      20

-------
    4.1.3    Documentation Required for Each Sample Collected:

        •    An individual and unique sample identifier and date of collection.  This
            must be recorded on the sample container in addition to the field  data
            records (notebook or form).
        •    Name of person  collecting the sample and specific sampling location
            data from which the sample was removed.
4.2 QC Samples

    4.2.1    Blank Samples.  Blank samples should be periodically collected at
            random throughout the sampling day at each sampling site. Two types
            of blank samples should be collected: field blanks and QC blanks.  Both
            these blanks are collected in the same manner; however, they are used
            for different purposes.

        4.2.1.1  Field blanks.  Field blank samples are used to identify any potential
                systematic lead contamination present in the wipe and during the
                handling of samples during field collection and laboratory analysis
                activities.  Field blanks should be collected in the same manner as
                used to collect field samples with the exception that no surface is
                wiped. Each wipe designated as a field blank should be removed
                from the bulk pack, folded to match the  field samples, and  placed
                into a labeled sample collection container.

                Each field blank must be labeled with its own unique identifier.  The
                identifier for all blanks should be similar to other field samples to
                mask the identify of the blank from the laboratory (i.e., blanks can
                then be submitted in  a blind manner to the laboratory). It is
                recommended that field blanks be collected at a frequency of 1 per
                20 field samples. At a minimum, three should be collected at each
                sampling site for each new pack of bulk wipes used for sample
                collection (i.e., one near the beginning of the sampling period at the
                site, one in the middle, and one near the end).  Field blank lead
                results should not exceed 20 u,g/sample. Lead  results above this
                value should trigger an investigation into the potential cause  and
                resampling of samples associated with the field blank may have to
                be undertaken.  Large blank lead values can often be sporadic and
                not systematic; therefore, blank correction  of field sample results
                using field blank data is not recommended.

        4.2.1.2  QC blanks.  QC blank samples are used for preparation of blind
                reference material samples described in subsection  4.2.2.  QC
                blanks should be collected in exactly the same manner as  described
                                      21

-------
                for field blanks. Each QC blank must be labeled with its own unique
                identifier.  The identifier for all blanks should be similar to other field
                samples to mask the identify of the blank from the laboratory (i.e.,
                reference  materials prepared from the blanks can then be submitted
                in a blind  manner to the  laboratory).  It is recommended that QC
                blanks be collected at a  frequency of 1 per 20 field samples. At a
                minimum,  two should be collected at each sampling site (an extra
                should be collected to assure sufficient QC blanks are available in
                case problems are experienced during preparation of blind reference
                material samples).

    4.2.2    Blind  Reference Material Samples.  Reference materials should be
            periodically submitted to the laboratory for analysis as a check on
            adherence to proper laboratory sample preparation and instrumental
            analysis methods.  Prepare a blind reference material by placing an
            accurately weighed portion (0.3000-1.0000 gram) of a reference material
            into a wipe (QC blank). The  wipe should be folded to the same degree
            as the field samples. Place the reference material containing wipe inside
            a labeled sample collection container.  Include a dummy sampling area
            on the label to disguise the identity of the blind reference material. The
            weight of reference material should be chosen to produce a blind
            reference material sample that will produce a lead level between 200  and
            1000  |a.g/sample.  It is recommended that the frequency of these QC
            samples, submitted to the laboratory for lead determinations, be at least
            1 per  20 field samples.  Reference materials from MIST6, such as SRMs
            2709, 2711, and 2704, are readily available and can be used for
            preparing blind reference materials.  Other sources of materials with
            known lead levels, such as performance samples from the ELPAT7
            program, also  may be used to prepare blind reference materials.
4.3 Contamination Avoidance

The following work practices should be followed to prevent cross-contamination of
samples:

        •    Avoid disturbing and tracking dust from one location to another by:
                identifying and clearly marking all sampling locations upon arrival at
                the sampling site,
                avoiding walking through or over any of the marked sampling
                location areas, and
                instructing field team members to pull on new disposable shoe
                covers upon each entry into the building.
        •    Change gloves frequently. Collection of each new sample must be
            conducted with a new pair of gloves.
                                      22

-------
            Clean sampling equipment and measuring tapes frequently with wipes.
            Inspect all sampling equipment for cleanliness prior to collection of each
            sample. Always clean suspect equipment if in doubt.
            Do not open sample collection containers until needed to collect each
            sample.
            When using bulk packed wipes, at each sampling location, discard the
            first two wipes pulled from wipe container.
5.0 Glossary

    5.1  Digestion. Sample preparation process that solubilizes lead present in the
        sample.  The digestion process produces an  acidified, aqueous solution called
        the digestate.  A lead determination is made  on the digestate during an
        instrumental measurement process.

    5.2  Field Blank. See subsection 4.2.1.

    5.3  Field Data. Any information collected at the sampling site.

    5.4  Field Sample. Physical material taken from the sampling site that are targeted
        for lead determination.

    5.5  Reference Material. Material of known composition containing a known
        amount of lead. These materials have typically been subjected to a large
        amount of lead determinations to develop a lead result known to a high
        degree of confidence.

    5.6  Sample Collection Container. Container for holding and transporting the
        samples from the field to the laboratory.  The internal volume of the container
        must be sufficient to hold the entire collected sample.

    5.7  Sampling Location. Specified area  within a sampling site that is subjected to
        sample collection.  Multiple sampling locations  are commonly designated for a
        single sampling site.  An example would be at the bottom of a specific slide in
        a specific playground area.

    5.8  Sampling Site.  Local geographical  area that contains the sampling locations.
        A sampling site is generally limited to an area that can be easily covered on
        foot.  An example would be John Smith's house at 3102 Nowhere Avenue,
        Detroit, Ml.
                                      23

-------
  5.9  Sample Preparation. Process used to ready a sample received from the field
      for lead determination using instrumental measurement methods.  The
      process is dependent on the sample type and can include a large number of
      steps such as homogenization, drying, splitting, weighing, digestion, dilution to
      a final known volume, and filtering.

5.10  QC Blank. See subsection 4.2.2.
                                    24

-------
D.  Protocol for Collection of Dust Samples for Lead Determination
    Using Vacuum Sampling

1.0 Introduction

This protocol provides for the collection of settled dust samples from surfaces using
vacuum methods. The protocol is suitable for the collection of settled dust samples
from both hard or smooth and highly textured surfaces, such as brickwork and rough
concrete, and soft, fibrous surfaces, such as upholstery and carpeting.

Procedures presented in this protocol are intended to provide a method for collection
of dust from surfaces that can not be sampled using wipe collection methods. In
addition, these procedures are written to utilize equipment that is readily available and
in common use for  other environmental sampling applications (i.e., air paniculate
sample collection).

Due to the flow dynamics inherent in the vacuum method, results for vacuum dust
samples are not likely to reflect the total dust contained within the sampling area. This
protocol generally will have a collection bias toward smaller, less dense, dust particles.
However, the protocol, if performed as written, will generate dust lead data that will be
consistent and comparable between operators performing the method.  This protocol
can be used to produce samples  for lead determination results in both loading dig/ft2)
and concentration (u,g/g).  It is recommended, however, that it not be used for the
generation of concentration results due to particle size collection bias and potential
errors intrinsic to processing and  handling preweighed filters (or entire filter cassettes),
which are  required to determine total collected sample weight.  Even though it is not
normally recommended, this protocol includes procedures for generation of total
collected sample weight.

Other vacuum sampling methods  that utilize less common equipment, such as cyclone
sample collectors,  may be useful for collection of settled dust, particularly with respect
to generation of more quantitative dust  lead concentration results.
2.0 Equipment and Supplies

2.1 Sampling Equipment

    2.1.1    Air-sampling pump. A portable, battery-powered air pump that is capable
            of a flow rate of 2.5 L/min through a filter cassette equipped with the
            nozzle specified in subsection 2.1.2.  Inlet of the pump must be fitted
            with a nipple to accept the tubing sized to fit tightly on the outlet side of a
            filter cassette.
                                      25

-------
2.1.2   Collection nozzle. A piece of stainless steel or carbon-impregnated
        plastic machined or molded on each end as follows: one machined or
        molded end to accept the tubing sized to fit tightly on the inlet side of a
        filter cassette; the other machined or molded to form a thin rectangular
        opening of Va inch by 3/(* inch.

2.1.3   Disposable shoe covers (optional). See subsection 4.3.

2.1.4   Filter cassette. 37-mm filter cassette, preloaded with 0.8-u.m,  pore-size
        Mixed-Cellulose Ester Filters (MCEF) and backup support pad.  If lead
        concentration results are to be determined and reported, then a special
        handling of these cassettes is required, as described in  subsections 3.3.1
        and 3.3.2.

2.1.5   Masking tape,  used for holding down sampling templates and marking
        sampling locations.

2.1.6   Soap bubble air flow meter or calibrated rotameter. equipped with inlet
        and outlet fittings sized to fit tubing used to connect the filter  cassette to
        the air-sampling pump.

2.1.7   Sampling templates, one ft2 inside area reusable aluminum or plastic, or
        disposable cardboard or plastic template.  A variety of shapes are
        recommended for use  in variable field situations such  as:  square,
        rectangular, square "U" shaped, rectangular "U" shaped, and "L." All
        templates must have accurately known inside  dimensions.  Templates
        should  be thin (less than Ve inch), and be  capable of lying flat on a flat
        surface.

2.1.8   Secondary sample  collection container, resealable plastic bags for
        holding and transporting the filter cassettes.

2.1.9   Steel or plastic measuring tape or ruler, divisions  to at least Vie inch.

2.1.10  Tubing, plastic, flexible tubing sized to fit tightly on both the inlet and
        outlet of a  filter cassette and the  inlet of the air-sampling pump.

2.1.11  Wipe, disposable towelette moistened with a wetting agent.   Wipe
        brands or sources should contain insignificant background lead levels.
        Laboratory analysis on replicate blank wipes should be  used  to
        determine  background lead levels prior to use in the field.   Background
        lead levels less than 10 u.g per wipe are considered insignificant for most
        dust-sampling activities.  It is recommended to avoid brands of wipes
        that contain aloe because wipes containing aloe have been found to
        contain higher background lead levels.
                                   26

-------
2.2 General Supplies

    2.2.1    Field notebooks, bound with individually numbered pages, see
            subsection 4.1.

    2.2.2    Indelible ink marker, black or blue.

    2.2.3    Ink pens, black or blue.

    2.2.4    Packaging tape, used for sealing shipping containers.

    2.2.5    Plastic bags, trash bags with ties.

    2.2.6    Plastic gloves, powderless.  Gloves with powder should not be used to
            avoid potential contamination of samples from powder material.

    2.2.7    Preprinted field forms,  preprinted with sufficient entry lines to address
            documentation needs presented in subsection  4.1

    2.2.8    Shipping containers, cardboard or plastic for interim storage and
            shipment of sample collection containers.


3.0 Sampling Procedure

Two types of sampling procedures are presented. The first, Loading Only Vacuum
Collection, is intended for collection  of dust for lead loading determinations (i^g/ft2)
only.  The second, Collection on  Preweighed Media, is intended for collection of dust
for both lead loading (u.g/ft2) and  lead concentration (u.g/g) determinations.  The later
type has two options that differ in the methods used for determining the total collected
sample weight.


3.1 Calibration of Air-Sampling Pump

Regardless of  the type of the sampling procedure used (see subsection 3.2 or 3.3),
the air-sampling pump used for sample collection must be calibrated prior to sample
collection for any given day. The procedure for air pump calibration  is as follows:

    3.1.1    Label a filter cassette with an ink marker to distinguish it as one used for
            pump calibration (and not to be confused with or used for collection  of a
            field sample). Remove the inlet and outlet plugs and place them in a
            labeled, resealable  plastic bag.
                                       27

-------
    3.1.2    Attach the filter cassette to the air-sampling pump with a piece of flexible
            tubing.  Attach a collection nozzle to the inlet side of the filter cassette
            using a short section of tubing (less than Vz inch).

    3.1.3    Insert a soap bubble meter, calibrated rotameter, or other equivalent
            calibrated flow rate measuring device in-line between the air pump and
            the filter cassette equipped with a nozzle.

    3.1.4    Turn on the air pump and adjust the flow rate of the air-sampling pump
            (if possible) to achieve an air flow between 2.5-2.8 L/min. Replace the
            air-sampling pump if this flow rate cannot be reached. Document the
            calibration  in field data records (notebook or forms).

At the end of the sample collection day, the calibration must be verified. Perform the
verification in the same manner as indicated in steps 3.1.1, 3.1.2, 3.1.3, and 3.1.4
above.  Document the calibration verification in a field data records (notebook or
forms).  If the calibration verification fails to reproduce the minimum flow rate  of
2.5 L/min, then all samples collected during the day are questionable and should be
discarded.
3.2 Loading Only Vacuum Collection Procedure

The following procedure assumes that concentration results will not to be determined.
In addition, it assumes that the air-sampling pump has been warmed up, and the
calibration has been performed as described in subsection 3.1.

Following is a summary of this procedure:

    1.   Select a sampling location.
    2.   Mark the sampling location.
    3.   Perform first vacuuming: One direction, side-to-side.
    4.   Perform second vacuuming: One direction, top-to-bottom.
    5.   Perform third vacuuming: One direction, side-to-side, store the sample.

The detailed procedure  is as follows:

    3.2.1     Pull on a pair of clean, powderless, plastic gloves.

    3.2.2    Mark the area to be sampled using one of the following two procedures:

        3.2.2.1  Template Assisted Marking. Carefully place a clean template on the
        surface in manner  that minimizes disturbance  of settled dust at the location.
        Either tape or place a heavy object on the outside edge of the template to
        prevent  it from moving during sample collection.
                                       28

-------
    3.2.2.2  Manual Marking of Sampling Area.  Mark an outline of the sampling
    location using masking tape.  Care should be taken to minimize any disruption
    of dust within the sampling location.  For areas that are dirty or contain high
    dust levels, new tape may have to be applied more than once to get adhesion
    to the surface.  Discard any soiled tape in a trash bag.

3.2.3    Discard any gloves used to mark the area in a trash bag and  pull on a
         new pair of clean, powderless, plastic gloves.

3.2.4    If not prelabeled from prefield processing,  label a filter cassette with an
         ink marker. Remove the inlet and outlet plugs and place them into a
         labeled resealable plastic bag.  Attach the outlet  to the air-sampling
         pump with a piece of flexible tubing.  Attach collection nozzle  to the inlet
         side of the filter cassette using a short section of new tubing (less than
         Va inch). Always use a new section of tubing for the inlet side of the filter
         cassette.

3.2.5    First Vacuuming: One Direction. Side-to-Side:  With the air-sampling
         pump on, vacuum the selected sampling surface area, starting at either
         of the corners furthest from the operator (referred to as a far corner),
         using a slow side to side (left-to-right or right-to-left) sweeping motion
         while holding the collection nozzle at an angle of approximately 45° to
         the sampling surface. Avoid pressing down hard on the sampling
         surface during sample collection. Move the nozzle at a rate of
         approximately 2-4 inches per second. At the end of the first pass from
         one side to the other, carefully lift the collection nozzle and repeat the
         vacuuming sweep in the  same direction as the first, using a slightly
         closer overlapping pass.  Care must be taken to  avoid overloading of the
         filter cassette.  Repeat the  procedure until the entire sampling area has
         been covered using the one-direction, side-to-side sweeping motions.

         Overloading will result in  decreased air flow and  a reduction in sampling
         efficiency and increased  sampling bias toward smaller, less dense
         particles.  A drop of air flow of more than 10% is an indicator  of
         overloading. If overloading of samples becomes evident, reduce the
         sampling area to prevent filter overloading or  use multiple cassettes for
         collection  within the same sampling area.

3.2.6    Second Vacuuming: One Direction. Top-to-Bottom:  With the air-sampling
         pump on,  vacuum the selected sampling surface area, starting at a far
         corner,  using a slow top-to-bottom sweeping motion in the same manner
         as described in subsection  3.2.5.  Repeat the procedure until the entire
         sampling area has been  covered using the one-direction, top-to-bottom
         sweeping motions.
                                   29

-------
    3.2.7    Third Vacuuming: One Direction. Side-to-Side: With the air-sampling
            pump on, vacuum the selected sampling surface area, starting at a far
            corner, using the slow, one-direction, side-to-side sweeping motion
            described in subsection 3.2.5.   Repeat the procedure until the entire
            sampling area has been covered using the one-direction, side-to-side
            sweeping motions.

    3.2.8    Remove the filter cassette from the inlet and outlet tubing sections,
            replace the cassette plugs, and place the sample into a labeled
            resealable plastic bag. Using a tape measure, measure the dimensions
            of the sampled area to within Vie inch (or verify the dimensions of the
            template) and label the plastic bag containing the sample with sufficient
            information to uniquely identify the sample and the dimensions of the
            selected sampling area (with units such as inches).  Also record this
            information on a preprinted data form or in a field notebook.  Discard the
            used gloves in the trash bag.
3.3 Loading and Concentration Vacuum Collection Procedure

The collection procedure used for reporting both loading and concentration results is
the same collection procedure as described in subsection 3.2 with two exceptions.
First, a prefield and postfield, stabilization-weighing procedure is required to determine
the total sample weigh collected. These procedures are described below.  Second,  if
the Option 2 procedures listed below are used for weight determinations, care must  be
exerted during all handling of the sample cassettes to avoid inadvertent additions of
weight to the filter cassettes.  Option 2 always requires that the sample cassettes be
handled with  gloves and never with bare hands.

The overall collection procedure assumes that the air-sampling pump has been
warmed up and the calibration has been performed as described in subsection 3.1.
Following is a summary of the overall collection procedure:

    1.   Perform prefield stabilization and gravimetric procedures.
    2.   Select a sampling location.
    3.   Mark the sampling location.
    4.   Perform first vacuuming: One direction, side-to-side.
    5.   Perform second vacuuming:  One direction, top-to-bottom.
    6.   Perform third vacuuming: One direction, side-to-side, store the sample.
    7.   Perform postfield stabilization and gravimetric  procedures.

The two options available for determining a prefield and postfield sample media
weight, preweighed filter and  preweighed filter cassette, are presented below:
                                       30

-------
3.3.1    Prefield Stabilization and Gravimetric Procedure, Option 1—Preweighed
        Filter

        This procedure suffers from the lack of quantitative transfer of all dust
        clinging to the cassette during postfield processing. Therefore, this
        option is considered somewhat more qualitative than Option 2.  However,
        unlike Option 2, it is not susceptible to weight errors resulting from
        inadvertent touching or improper handling of the filter cassettes between
        pre- and postfield processing.

    3.3.1.1   Prefield Procedure.  The filter inside the cassette (not the backup
    support pad) must be weighed to constant weight prior to sample collection
    (prefield) at known temperature and humidity conditions (i.e., desiccated at
    room temperature).  This can be performed for preloaded filter cassettes as
    follows:

    a.   Pull on a  new pair of clean, powderless, plastic gloves.
    b.   Place a unique sample identifier on the outside of each cassette targeted
        for preweight generation using  indelible ink and allow to dry.
    c.   Using a clean screwdriver, separate the cassette rings that hold the filter
        in place.  Place the rings on a clean, dry area, such as a plastic bag or
        equivalent surface.
    d.   Using clean plastic tongs, lift the filter from the cassette and place it in  a
        clean, dry, labeled beaker, watch glass, or other equivalent labeled
        container.
    e.   Place the container with filter into a desiccator and allow the filter to
        stabilize to a constant  weight.  Periodically weigh and record the filter on
        a clean balance to determine weight stability.  (Record all weights to
        ±0.0001 g.) A constant weight for this protocol is one that does not
        change more than ±0.002 gram for repeated measurements (minimum  of
        2) taken over a  minimum of a 24-hour period.  Using clean plastic tongs,
        replace the filter back  into the cassette,  reassemble the cassette,
        reweigh the container, and record the empty container weight. The
        prefield filter weight is  the difference between the container plus filter
        weight and the container-only filter weight.
    f.   Place the preweighed  filter inside the sample cassette into a resealable
        plastic bag container for transport to the field.

    3.3.1.2  Postfield Procedure.  The filter and dust inside the cassette (not the
    backup support pad) must be weighed to constant weight prior to laboratory
    sample preparation (postfield) at the same known temperature and humidity
    conditions used for prefield processing.  This can be performed as follows:

    a.   Pull on a  new pair of clean, powderless, plastic gloves for each sample
        handled.
                                   31

-------
    b.   Place a unique sample identifier on the outside of a clean digestion
        vessel (usually a borosilicate glass beaker) using indelible ink and allow
        to dry. Tare the beaker (determine and record the weight) to ±0.0001 g.
    c.   Using a clean screwdriver, carefully separate the cassette rings that hold
        the filter in place while holding the cassette over the labeled beaker.
        Allow any dust contained  inside the cassette to fall into the beaker.
        Using clean plastic tongs, carefully lift the filter from the cassette and
        drop it into the beaker.  Carefully tap any visible dust clinging to the
        inside of the cassette into the beaker.
    d.   Place the beaker with filter into a desiccator and allow the filter to
        stabilize to a constant weight. Periodically weigh and record the weight
        of the container plus filter. (Record all weights to ±0.0001 g.)  A constant
        weight for this protocol is  one that does not change more than ±0.002
        gram for repeated measurements (minimum of 2) taken over  a minimum
        of a 24-hour period. The  postfield filter weight is the difference between
        the container plus filter and dust weight and the container-only weight.
        Due to the potential of the dust to have significant water absorption,
        stabilization times for postfield weighing is expected to be considerably
        longer than for prefield gravimetric.  It  is  recommended  (not required)
        that no initial weight data  be attempted until the sample has remained in
        the desiccator for at least 72 hours.
    e.   The entire sample plus filter in the beaker must be prepared for lead
        analysis.  The total sample weight for use in determining lead
        concentration is the difference between the postfield filter weight and the
        prefield filter weight.

3.3.2    Prefield Stabilization and  Gravimetric Procedure, Option 2—Preweighed
        Filter Cassette

        This procedure results in  a better quantitative transfer of all dust clinging
        to the cassette during postfield processing.  Therefore, this option is
        considered somewhat more quantitative than Option 1.  However, it is
        susceptible to weight errors resulting from inadvertent touching or
        improper handling of the filter cassettes between pre- and postfield
        processing.

    3.3.2.1   Prefield Procedure.  The entire filter cassette must be weighed to
    constant weight prior to sample collection (prefield) at known temperature and
    humidity conditions (i.e., desiccated at room temperature). All handling of the
    cassettes must be done with gloves.  This can be performed for preloaded
    filter cassettes as follows:

    a.   Pull on a new pair of clean, powderless,  plastic gloves.
    b.   Place a unique sample identifier on the outside of each cassette targeted
        for preweight generation using  indelible ink and allow to dry.
                                   32

-------
        c.    Remove the inlet and outlet plugs and place them into a labeled,
             resealable plastic bag.
        d.    Place the filter cassette into a desiccator in a manner that allows air to
             flow freely through the inlet and outlet holes.  Allow the filter cassette to
             stabilize to a constant weight. Record the weight of the entire filter
             cassette without plugs. (Record all weights to ±0.0001 g.)   A constant
             weight for this protocol is one that does not change more than
             ±0.002 grams for repeated measurements (a minimum  of 2) taken over a
             minimum of a 24-hour period.
        e.    Replace the inlet and outlet plugs and place the entire filter cassette with
             plugs into a labeled resealable plastic bag for transport to the field.

        3.3.2.2   Postfield Procedure.  The filter cassette with dust (without plugs)
        must be weighed to constant weight prior to laboratory sample preparation
        (postfield) at the same known temperature and humidity conditions used for
        prefield processing. This can be performed as follows:

        a.    Pull  on a new pair of clean, powderless, plastic gloves.
        b.    Remove the inlet and outlet plugs and place them back into the original
             labeled,  resealable plastic bag.
        c.    Place the filter cassette into a desiccator in a manner that allows air to
             flow freely through the inlet and outlet holes and that does not allow any
             spillage of dust out the holes. Allow the filter cassette to stabilize to a
             constant weight.  Record the weight of the entire filter cassette plus dust
             without plugs. (Record all weights to ±0.0001 g.)  A constant weight for
             this  protocol is one that does not change more than ±0.002 grams for
             repeated measurements (minimum of 2) taken over a minimum  of a
             24-hour period. It is recommended (not required) that no initial weight
             data be attempted until the sample has remained in the desiccator for at
             least 72  hours.
        d.    The  contents of the filter cassette should be prepared for lead analysis.
             A quantitative transfer procedure that utilizes the backup support pad for
             wiping dust out of the inside of the cassette combined with rinsing out
             the cassette with dilute acid can be used to transfer the entire sample to
             the digestion vessel. The total sample weight for use in determining lead
             concentration is the difference between the postfield filter cassette weight
             and  the prefield filter cassette weight.
4.0 Quality Control

Adherence to quality control (QC) procedures is an important part of field sample
collection.  QC procedures, including documentation requirements, field QC samples,
reference material check samples, and contamination avoidance are presented in this
section.
                                       33

-------
4.1 Documentation

All field data related to sample collection must be documented.  A field notebook or
sample log form can be used to record field collection data. It is recommended to
utilize both types of documentation records (field notebooks and preprinted sample log
forms) for assuring collection of all relevant field data. Field data entries on
documentation records must adhere to the following requirements:

    4.1.1    General Documentation Requirements:

        •   All entries must be made using ink.
        •   Each page (notebook or form) must include the name of the person
            making the entries and the date of entries found on the page.
        •   Any  entry errors must be corrected by using only a single line through
            the incorrect entry (no scratch outs) and marked with the initials of the
            person making the correction and the date of correction.
        •   An initial page that correlates  initials to a specific name must be
            generated and maintained with field data records to trace any initials
            used in notebooks and on data forms.

    4.1.2   Specific Sampling Site Documentation Requirements:

        •   General sampling site description.
        •   Project or client name, address, and city/state location.
        •   Information as to what specific collection protocol was used.
        •   Information as to the use of interim storage and sample shipment
            mechanisms.
        •   Prefield weight data including  stabilization conditions for filter cassette.
        •   Postfield weight data including stabilization conditions for filter cassette.

    4.1.3   Documentation Required for Each Sample Collected:

        •   An individual and unique sample identifier and date of collection. This
            must be recorded on the sample container in addition to the field data
            records (notebook or form).
        •   Name of person collecting the sample and specific sampling location
            data from which the sample was removed.
4.2 QC Samples

    4.2.1    Blank Samples.  Blank samples should be periodically collected
            (designated) throughout the sampling day at each sampling site. Two
            types of blank samples should be collected; field blanks and QC blanks.
                                      34

-------
         Both these blanks are collected in the same manner; however, they are
         used for different purposes.

    4.2.1.1   Field blanks.  Field blank samples are used to identify any potential
             systematic lead contamination present in the filter cassette and
             handling of samples during field collection  and laboratory analysis
             activities.  Field blanks should be collected in the same manner
             used to collect field samples with the exception that no air is drawn
             through the filter cassette.  Each cassette  designated as a field
             blank should be removed from the plastic  bag, inlet and outlet caps
             pulled off, the tubing and sampling nozzle  attached, and then this
             procedure  is reversed. The vacuum pump is not turned on.

             Each field  blank must be labeled with its own unique identifier.  The
             identifier for all blanks should be similar to other field samples to
             mask the identify of the blank from the  laboratory (i.e., blanks can
             then be submitted in a blind manner to the laboratory). It  is
             recommended that field blanks be collected (or designated) at a
             frequency of 1 per 20 field samples. At a  minimum, one field blank
             should be collected at each sampling site. Field blank lead results
             should not exceed 20 [ig/sample. Lead results above this value
             should trigger an investigation into the potential cause of the
             problem and resampling of samples associated with the field blank
             may have to be undertaken.  Large blank  lead values can often be a
             sporadic and not systematic; therefore, blank correction of field
             sample results using field  blank data is not recommended.

    4.2.1.2   QC blanks. QC blank samples are  used for preparation of blind
             reference material samples described in subsection 4.2.2. QC
             blanks should be collected in exactly the same manner as described
             for field blanks. Each QC blank must be labeled with  its own unique
             identifier. The identifier for all blanks should be similar to other field
             samples to mask the identify of the  blank from the laboratory (i.e.,
             blind reference materials prepared from the blanks can then be
             submitted in a blind manner to the laboratory).  It is recommended
             that QC  blanks be collected (or designated) at a frequency of 1 per
             20 field samples.  At a minimum, two should be collected at each
             sampling site  (an extra should be collected to assure sufficient QC
             blanks are available in case problems are  experienced during
             preparation of blind reference material samples).

4.2.2     Blind Reference Material Samples.  Reference materials should  be
         periodically submitted to the laboratory for analysis as a check on
         adherence to proper laboratory sample preparation and instrumental
         analysis methods.   Prepare a blind reference material by placing an
                                   35

-------
            accurately weighed portion (0.3000-1.0000 gram) of a reference material
            into a blank filter cassette (a collected QC blank).  Place the cassette
            with weighed reference material inside a labeled, sample collection
            container. Include a dummy sampling area on the label to help disguise
            the blind reference material sample.  The weight of reference material
            should be chosen to produce a blind reference material sample that will
            produce a lead level between 200 and 1000 (ig/sample.  It is
            recommended that the frequency of these QC samples submitted to the
            laboratory for lead determinations be at least 1 per 20 field samples.
            Reference materials from NIST6, such as SRMs 2709,  2711, and 2704,
            are readily available and can be used for preparing blind reference
            materials. Other sources of materials with known lead levels, such as
            performance evaluation materials from the ELPAT7 program, also may
            be used to prepare blind reference materials.
4.3 Contamination Avoidance

The following work practices should be followed to prevent cross-contamination of
samples:

        •    Avoid disturbing and tracking dust from one location to another by:
                identifying and clearly marking all sampling locations upon arrival at
                the sampling site,
                avoiding walking through or over any of the marked sampling
                location areas, and
                instructing field team members to pull on new disposable shoe
                covers upon each entry into the building.
        •    Use a new pair of powderless gloves at each sampling  location.
        •    Inspect all sampling equipment for cleanliness prior to collection of each
            sample.  Always clean suspect equipment if in doubt.
        •    Do not open sample collection containers until needed to collect each
            sample.
        •    When using bulk packed wipes, at each sampling location, discard the
            first two wipes pulled from the wipe container.
5.0 Glossary

    5.1  Digestion. Sample preparation process that solubilizes lead present in the
        sample. The digestion process produces an acidified, aqueous solution called
        the digestate. A lead determination is made on the digestate during an
        instrumental measurement process.

    5.2  Field Blank. See subsection 4.2.1.


                                       36

-------
  5.3 Field Data. Any information collected at the sampling site.

  5.4 Field Sample. Physical material taken from the sampling site that is targeted
      for lead determination.

  5.5 Reference Material. Material of known composition containing a known
      amount of lead. These materials have typically been subjected to a large
      number of lead determinations to develop a lead result known to a high
      degree of confidence.

  5.6 Sample Collection Container. Container for holding and transporting the
      samples from the field to the laboratory.  The internal volume of the container
      must be sufficient to hold the entire collected sample.

  5.7 Sampling Location. Specified area within a sampling site that is subjected to
      sample collection.  Multiple sampling locations are commonly designated for a
      single sampling site.  An  example would be at the bottom of a specific slide in
      a specific playground area.

  5.8 Sampling Site.  Local geographical area that contains the sampling locations.
      A sampling site is generally limited to an area that can  be easily covered on
      foot.  An example would be John Smith's house at 3102 Nowhere Avenue,
      Detroit, Ml.

  5.9 Sample Preparation. Process used to ready a sample received from the field
      for lead determination using instrumental measurement methods.  The
      process  is dependent on  the sample type and can include a large  number of
      steps such as homogenization, drying, splitting, weighing, digestion,  dilution to
      a final known volume, and filtering.

5.10  QC Blank. See subsection 4.2.2.
                                    37

-------
E.  References

1    ASTM ES 29-94, Practice for the Field Collection of Soil Samples for Lead
    Determination by Atomic Spectrometry Techniques, American Society of Testing
    and Materials, Committee E-6, Performance of Buildings, ASTM
    PCN:03-506194-10, 1994.

2    ASTM ES 30-94, Practice for the Field Collection f Settled Dust Samples Using
    Wipe Sampling  Methods for Lead Determination by Atomic Spectrometry
    Techniques, American Society of Testing and Materials, Committee E-6,
    Performance of Buildings, ASTM PCN:03-506194-10, 1994.

3    Environmental Protection Agency, Office of Solid Waste, Test Methods for
    Evaluating Solid Waste Physical/Chemical Methods. U.S. EPA SW 846, Third
    Edition, Proposed Update II.

4    ASTM ES 36-94, Practice for Hot Plate Digestion of Dust Samples for
    Determination of Lead by Atomic Spectrometry, American Society of Testing and
    Materials, Committee E-6, Performance of Buildings, ASTM PCN:03-506194-10,
    1994.

5    ASTM E 1613-94, Test Method for Analysis for Lead by Inductively Coupled
    Plasma Atomic  Emission Spectrometry (ICP-AES), Flame Atomic Absorption
    Spectrometry (FAAS), or Graphite Furnace Atomic Absorption Spectrometry
    (GFAAS) Technique, American Society of Testing and Materials, Committee E-6,
    Performance of Buildings ASTM PCP:03-506194-10, 1994.

6    National Institute of Standards and Technology, U.S. Department of Commerce,
    Technology Administration, Standard Reference Materials Catalog.  1993-94.

7    Environmental Lead Proficiency Analytical Testing (ELPAT) Program, performance
    samples remaining after performance testing rounds, ELPAT Coordinator,
    American Industrial Hygiene Association (AIHA), 2700 Prospect Avenue, Suite
    #250, Fairfax, VA 22031, (703) 849-8888.
                                     38

-------