SW84632


   FINAL  (PROMULGATED) UPDATES II AND  MA
                        Cover Sheet

      THIS PACKET CONTAINS NEW AND REVISED MATERIAL
                       FOR INCLUSION IN:
         TEST METHODS FOR EVALUATING SOLID WASTE
                 PHYSICAL/CHEMICAL METHODS
                     (SW-846) THIRD EDITION
Contents:

 1.  Cover sheet.  (What you are currently reading)

 2.  Instructions.  This section explains how to put together your SW-846
     manual.
      A. Instructions for New Subscribers.
      B. Instructions for Previous Subscribers.

 3.  Method  Status Table.  The Method Status Table is  a sequentially
     numbered listing of all SW-846 methods and their current status.

 4.  Revised Update II Table of Contents.  The Table of Contents (dated
     September 1994) lists all of the methods (Third Edition, Update I, and
     Updates II and IIA) in  the order in which they should appear in the
     manual.

 5.  Revised Chapter Two: Choosing the Right Method

 6.  Revised Chapter Three and new/revised methods for metals analyses.

 7.  Revised Chapter Four and new/revised methods for organic analyses.

 8.  Revised  Chapter  Five and  new/revised  methods for  miscellaneous
     analyses.

 9.  Revised Chapter Six and new/revised methods for properties analyses.

 10.  Revised Chapter Seven:  Introduction and Regulatory Definitions

 11.  Revised Chapter Eight (Revised section separation sheets only)

-------
                            INSTRUCTIONS

SW-846 is a "living" document that changes when new data and advances in
analytical techniques are incorporated into the manual as new or revised
methods.  Periodically, the Agency issues these methods as updates to  the
manual.  To date, the Agency has issued Final Updates I, II, and IIA. These
instructions include directions on getting the basic manual up-to-date and
incorporating Final Updates  II and IIA into your SW-846. The Agency will
release additional proposed and final updates in the future. New instructions,
to supersede these, will be included with each of those updates. However, in
general,  final  updates  should  always  be incorporated  into SW-846  in
chronological order (e.g. Update I should be incorporated before Update  II).

If you have any difficulty with these directions, you may telephone the Methods
Information Communication  Exchange (MICE) at 703-821-4789 for help.  If
you have questions  concerning your SW-846 U.S. Government Printing Office
(GPO) subscription, you should telephone the GPO at 202-512-2303.  If you
did not purchase your SW-846 from the GPO, the GPO will not be able to help
you.
FINAL UPDATE IIA;  Final Update IIA contains only one method, Method
4010, dated August 1993.  This method was promulgated on January 4, 1994
(59 FR 458).  It should be inserted into the manual according to the location
specified in the Final Update II Table of Contents (dated September 1994).
FINAL  UPDATE II;   Final  Update II has been  promulgated and is  now
officially part of SW-846. These instructions for insertion of Final Update II
are divided into two (2) sections: Section A - Instructions for New Subscribers
and Section B - Instructions for Previous Subscribers.

    New subscribers are defined as individuals who have recently (6-8 weeks)  placed an
    order with the GPO and have received new copies of the 4 (four) volume set of the
    Third Edition, a copy of Final Update I, and a copy of Final Updates II and IIA.

    Previous subscribers are defined as individuals that have received  copies of the Third
    Edition and other SW-846 Updates (including proposed Updates) in the past and have
    just  received their Final Update II and IIA package in the mail.
Update II and IIA               Instructions - 1                          Final

-------
Please use the following  instructions  for  new subscribers  or  previous       V
subscribers in sequence to piece together your new SW-846 manual.
A. INSTRUCTIONS FOR NEW SUBSCRIBERS

  i.  If you have not already done so, open the packages that contain the Third Edition of
    SW-846.  The Third Edition should include 4 (four) volumes of material (i.e. Volumes
    IA, IB, 1C, and II) and will be dated "September 1986" in the lower right hand corner
    of each page.  Four 3-ring binders (one binder for each volume) and a set of tabs
    should also be  included.  You  should place  each volume  of  material  in  the
    appropriately labeled 3-ring binder and insert the tabs. Check the Table of Contents
    (dated September 1986) if you have any questions about the order of the methods or
    about which volume the methods should be inserted into.

    You will be missing some methods from the Third Edition since any Third Edition
    September 1986 material, that was superseded by Final Update  I July 1992 material.
    has already been removed from your copy of the Third Edition.
 ii.  If you have not already done so, open the package that contains Final Update I. Final
    Update I should be a single package printed on white paper with the date "July 1992"
    in the lower right hand corner of each page. This package contains new methods and
    revised methods. In order to have a complete SW-846 manual, you should insert the
    new and revised July 1992 material using the Table of Contents (dated July  1992) at
    the front of Final  Update I to identify the correct location for  each  chapter and
    method.

    Since you are a new subscriber to SW-846, you need not be concerned about the
    removal or replacement of the previous version of Update I, as discussed in item (A)
    of the Final Update I instructions. Again, any Third Edition September 1986 material
    that was superseded by Final Update I July 1992 material, has already been removed
    from your copy of the Third Edition. For example, your copy of the Third Edition does
    not contain a copy of the September 1986 version of Chapter  One because it was
    superseded by the July 1992 revision of Chapter One contained in your Final Update
    I package.

    Final Update I also includes copies of September 1986 "replacement methods" which
    are included with your copy of Final Update I and are discussed in item (E) in the
    Final Update I instructions.  You should not insert the replacement methods!  The
    replacement methods were sent to subscribers before final Update II was released.
Update II and IIA                 Instructions - 2                             Final

-------
    The Disclaimer and Chapter One at the front of Update I should also be photocopied
    3 times and inserted at the front of volumes IB, 1C, and II in order to complete the
    manual.

         Note:   Update I does not contain any changes to Volume II other than the
         insertion of the Disclaimer and Chapter One.  Also, some methods will have an
         "A" after the method number. The "A" methods have been revised once.
iii.  Finally, open the package labeled Final Updates II and IIA. Final Updates II and IIA
    should be a single package printed on white paper. Update II has the date "September
    1994" in the lower right hand corner of each page. Update IIA (Method 4010) has the
    date "August 1993" in the lower right hand corner of each page.  This package contains
    new methods and revised  methods.  In order to have a complete SW-846 manual, you
    should insert the new methods and use the revised September 1994 methods to replace
    older Third Edition and Final Update I methods that are out of date. Use the Table
    of Contents (September 1994) at the front of Final Update II  to identify the correct
    location for each chapter  and  method.

    The Abstract and Table of Contents at the front of Final Update II should  also be
    photocopied 3 times and  inserted at the front  of volumes IB,  1C, and II in order to
    complete the manual.

    Please Note:

       •  Update II does not contain any changes to Volume II other than the insertion of
         the Abstract and Table of Contents.

       •  Some methods will have an "A" or a "B" after the method number.  The "A"
         methods have been revised once.  The "B" methods have been revised twice.

       •  Methods 5100, 5110, and 9200A were included in the  Proposed Update II
         (November 1992) package but are not included in the Final  Update II (September
         1994) package.  The final Federal Register Rule for Update II explains why these
         methods were not finalized (promulgated).
Update II and IIA                 Instructions - 3                             Final

-------
B. INSTRUCTIONS FOR PREVIOUS SUBSCRIBERS

 i.  Background Information:  A number of SW-846 update packages have been released
    to the public since the original Third Edition was released. The number and labels on
    these packages  can be confusing.  The following table titled "A Brief History of the
    SW-846 Third Edition and Updates" has been provided as an aid. Currently finalized
    (promulgated) methods have been printed in bold.  An individual or organization that
    has held an SW-846 GPO subscription for several years may have received copies of
    any or all of the following documents:
A BRIEF HISTORY OF THE SW-846 THIRD EDITION AND UPDATES
Package
Third Edition
Proposed Update I
Final Update I
(Accidently Released)
Proposed Update II
(Accidently Released)
Final Update I
Proposed Update II
Proposed Update IIA*
(Available by request only.)
Final Update HA* (Included
with Final Update II.)
Final Update II
Date Listed on Methods
September 1986
December 1987
November 1990
November 1990
July 1992
November 1992
October 1992
August 1993
September 1994
Color of Paper
White
Green
White
Blue
White
Yellow
White
White
White
Status of Package
Finalized (Promulgated)
Obsolete
Obsolete! Never formally
finalized.
Obsolete! Never formally
proposed.
Finalized (Promulgated)
Obsolete
Obsolete
Finalized (Promulgated)
Finalized (Promulgated)
    * Contains only Method 4010.

 ii.  In order to begin updating the manual it is important to establish exactly what is
    currently contained in the manual that you have.  If the manual has been properly
    updated, the ONLY white pages in the document should be dated September 1986
    (Third Edition) and July 1992 (Final Update I). Remove and discard (or archive) any
    white pages from your manual that have any date other than September 1986 and July
    1992.

    There may also be yellow pages dated September 1992 (Proposed Update II) inserted
    in the manual. Remove and discard all yellow pages or other colored pages (green or
    blue) from the manual.  Some  individuals may have chosen to keep their copy of
    Proposed Update II in a separate binder and removal will not be necessary.
U ?date II and IIA
Instructions - 4
Final

-------
iii.  Open the package labeled Final Updates II and IIA.  Final Updates II and IIA should
    be a single package printed on white paper.  Update II has the date "September 1994"
    in the lower right hand corner of each page.  Update IIA (Method 4010) has the date
    "August 1993" in the lower right hand corner of each page.  This package contains new
    methods and revised methods.  In order to have a complete SW-846 manual, you
    should insert the new methods and use the revised September 1994 methods to replace
    older Third Edition and Final Update I methods that are out of date. Use the Table
    of Contents  (September 1994) at  the front of Final Update II to identify  the correct
    location for  each chapter and method.

    The Abstract and Table of Contents at the front of Final Update II should also be
    photocopied 3 times and inserted at the front of volumes IB, 1C, and II  in order to
    complete the manual.

    Please Note:

       •  Update II does not contain any changes to Volume II other than the insertion of
         the Abstract and Table of Contents.

       •  Some methods will have an "A" or a "B" after the method number.  The "A"
         methods have been revised once. The "B" methods have been revised twice.
                 dit,>  cu.n 2-      o, 5
       •  Methods 5100, 5110,  and 9200A were included  in the  Proposed  Update II
         (November 1992) package but are not included in the Final Update II (September
         1994) package.  The final Federal Register Rule for Update II explains why these
         methods were not finalized (promulgated).
Update II and IIA                 Instructions - 5                             Final

-------
             METHOD STATUS TABLE
SW-846, THIRD EDITION, UPDATES I, II, AND HA
                 September 1994
      Use this table as a reference  guide to identify the
      promulgation status of SW-846 methods.

      The methods in  this table are  listed sequentially  by
      number.

      This table should not be used as a Table of Contents for
      SW-846. Refer to the Table of Contents found in Final
      Update II (dated September 1994) for the order in which
      the methods appear in SW-846.
                                          prot«*ion Agency

-------
SW-846 METHOD STATUS TABLE
      September 1994
NETH NO.
THIRD ED
DATED
9/86
0010
0020
0030
1010
1020
1110
1310
"

HETH NO.
FINAL
UPDATE I
DATED
7/92
~ "*
~ ~
~ ~
_ ••
1020A
~ ~
1310A
1311

NETH NO.
FINAL
UPDT. II
DATED
9/94
"
"™ ~
~ ~
"" ~
— —
_ _

™* ™
1312
NETHOD TITLE
Modified Method 5
Sampling Train
Source Assessment
Sampling System
(SASS)
Volatile Organic
Sampling Train
Pensky-Martens
Closed-Cup Method
for Determining
Ignitability
Setaflash Closed-Cup
Method for
Determining
Ignitability
Corrosivity Toward
Steel
Extraction Procedure
(EP) Toxicity Test
Method and
Structural Integrity
Test
Toxicity
Characteristic
Leaching Procedure
Synthetic
Precipitation
Leaching Procedure
SU-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol II
Chap 10
Vol II
Chap 10
Vol II
Chap 10
Vol 1C
Chap 8
Sec 8.1
Vol 1C
Chap 8
Sec 8.1
Vol 1C
Chap 8
Sec 8.2
Vol 1C
Chap 8
Sec 8.4
Vol 1C
Chap 8
Sec 8.4
Vol 1C
Chap 6
CURRENT
PROMUL-
GATED
METHOD
0010
Rev 0
9/86
0020
Rev 0
9/86
0030
Rev 0
9/86
1010
Rev 0
9/86
1020A
Rev 1
7/92
1110
Rev 0
9/86
1310A
Rev 1
7/92
1311
Rev 0
7/92
1312
Rev 0
9/94

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
1320
1330
3005
3010
	
3020
3040
3050
NETH NO.
FINAL
UPDATE I
DATED
7/92
"
1330A
3005A
3010A
•* —
3020A
~ —
3050A
NETH NO.
FINAL
UPDT. II
DATED
9/94
"
— ~


3015

~" ~*
"
METHOD TITLE
Multiple Extraction
Procedure
Extraction Procedure
for Oily Wastes
Acid Digestion of
Waters for Total
Recoverable or
Dissolved Metals for
Analysis by FLAA or
ICP Spectroscopy
Acid Digestion of
Aqueous Samples and
Extracts for Total
Metals for Analysis
by FLAA or ICP
Spectroscopy
Microwave Assisted
Acid Digestion of
Aqueous Samples and
Extracts
Acid Digestion of
Aqueous Samples and
Extracts for Total
Metals for Analysis
by GFAA Spectroscopy
Dissolution
Procedure for Oils,
Greases, or Waxes
Acid Digestion of
Sediments, Sludges,
and Soils
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
CURRENT
PROMUL-
GATED
METHOD
1320
Rev 0
9/86
1330A
Rev 1
7/92
3005A
Rev 1
7/92
3010A
Rev 1
7/92
3015
Rev 0
9/94
3020A
Rev 1
7/92
3040
Rev 0
9/86
3050A
Rev 1
7/92

-------
SW-846 HETHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
~ ""
3500
3510
3520
3540
~ "•
3550
3580
3600
METH NO.
FINAL
UPDATE I
DATED
7/92
™* —
3500A
3510A
3520A
3540A
~ ~
— ~*
3580A
3600A
NETH NO.
FINAL
UPDT. II
DATED
9/94
3051
™ ~
3510B
3520B
3540B
3541
3550A
"
3600B
HETHOD TITLE
Microwave Assisted
Acid Digestion of
Sediments, Sludges,
Soils, and Oils
Organic Extraction
and Sample
Preparation
Separatory Funnel
Liquid-Liquid
Extraction
Continuous Liquid-
Liquid Extraction
Soxhlet Extraction
Automated Soxhlet
Extraction
Ultrasonic Extrac-
tion
Waste Dilution
Cleanup
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.2
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.2
CURRENT
PROMUL-
GATED
NETHOD
3051
Rev 0
9/94
3500A
Rev 1
7/92
3510B
Rev 2
9/94
3520B
Rev 2
9/94
3540B
Rev 2
9/94
3541
Rev 0
9/94
3550A
Rev 1
9/94
3580A
Rev 1
7/92
3600B
Rev 2
9/94

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
3610
3611
3620
3630
3640
3650
3660
"
3810
METH NO.
FINAL
UPDATE I
DATED
7/92
3610A
3611A
3620A
3630A
"• ^
3650A
3660A
"
"
METH NO.
FINAL
UPDT. II
DATED
9/94
— —
~* ~
— ~
3630B
3640A
~ ~
~ ~
3665
"
METHOD TITLE
Alumina Column
Cleanup
Alumina Column
Cleanup and
Separation of
Petroleum Wastes
Florisil Column
Cleanup
Silica Gel Cleanup
Gel -Permeation
Cleanup
Acid-Base Partition
Cleanup
Sulfur Cleanup
Sulfuric
Acid/Permanganate
Cleanup
Headspace
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec 4.4
CURRENT
PROMUL-
GATED
METHOD
3610A
Rev 1
7/92
3611A
Rev 1
7/92
3620A
Rev 1
7/92
3630B
Rev 2
9/94
3640A
Rev 1
9/94
3650A
Rev 1
7/92
3660A
Rev 1
7/92
3665
Rev 0
9/94
3810
Rev 0
9/86

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
3820

5030
5040

"
6010
METH NO.
FINAL
UPDATE I
DATED
7/92
~ ~

5030A


"
6010A
METH NO.
FINAL
UPDT. II
DATED
9/94
~ ~
4010
(Update
IIA,
dated
8/93)
_ _
5040A
5041
5050

METHOD TITLE
Hexadecane
Extraction and
Screening of
Purgeable Organics
Screening for
Pentachlorophenol
by Immunoassay
Purge-and-Trap
Analysis of Sorbent
Cartridges from
Volatile Organic
Sampling Train
(VOST) : Gas
Chromatography/Mass
Spectrometry
Technique
Protocol for
Analysis of Sorbent
Cartridges from
Volatile Organic
Sampling Train
(VOST): Wide-bore
Capillary Column
Technique
Bomb Preparation
Method for Solid
Waste
Inductively Coupled
Plasma-Atomic
Emission
Spectroscopy
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec 4.4
Vol IB
Chap 4
Sec 4.4
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol 1C
Chap 5
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
3820
Rev 0
9/86
4010
Rev 0
8/93
5030A
Rev 1
7/92
5040A
Rev 1
9/94
5041
Rev 0
9/94
5050
Rev 0
9/94
6010A
Rev 1
7/92

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
— ~
7000
7020
7040
7041
7060
7061
~* ™*
7080
METH NO.
FINAL
UPDATE I
DATED
7/92
~ •"•
7000A
~ ~
~ ™
~ *~
~* ~~
7061A
"
— •*
METH NO.
FINAL
UPDT. II
DATED
9/94
6020
~ ~
~ "
— _
_ •*
7060A
"
7062
7080A
METHOD TITLE
Inductively Coupled
Plasma - Mass
Spectrometry
Atomic Absorption
Methods
Aluminum (Atomic
Absorption, Direct
Aspiration)
Antimony (Atomic
Absorption, Direct
Aspiration)
Antimony (Atomic
Absorption, Furnace
Technique)
Arsenic (Atomic
Absorption, Furnace
Technique)
Arsenic (Atomic
Absorption, Gaseous
Hydride)
Antimony and Arsenic
(Atomic Absorption,
Borohydride
Reduction)
Barium (Atomic
Absorption, Direct
Aspiration)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
6020
Rev 0
9/94
7000A
Rev 1
7/92
7020
Rev 0
9/86
7040
Rev 0
9/86
7041
Rev 0
9/86
7060A
Rev 1
9/94
7061A
Rev 1
7/92
7062
Rev 0
9/94
7080A
Rev 1
9/94

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
7420
7421
_ _.
7450
7460
"
7470
7471
7480
NETH NO.
FINAL
UPDATE I
DATED
7/92
_ _
— —
7430

_ _
7461
"
"

NETH NO.
FINAL
UPDT. II
DATED
9/94
"~ —
~ ~
— —
_ _
~ ~
"
7470A
7471A
"
METHOD TITLE
Lead (Atomic
Absorption, Direct
Aspiration)
Lead (Atomic
Absorption, Furnace
Technique)
Lithium (Atomic
Absorption, Direct
Aspiration)
Magnesium (Atomic
Absorption, Direct
Aspiration)
Manganese (Atomic
Absorption, Direct
Aspiration)
Manganese (Atomic
Absorption, Furnace
Technique)
Mercury in Liquid
Waste (Manual Cold-
Vapor Technique)
Mercury in Solid or
Semi sol id Waste
(Manual Cold-Vapor
Technique)
Molybdenum (Atomic
Absorption, Direct
Aspiration)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
7420
Rev 0
9/86
7421
Rev 0
9/86
7430
Rev 0
7/92
7450
Rev 0
9/86
7460
Rev 0
9/86
7461
Rev 0
7/92
7470A
Rev 1
9/94
7471A
Rev 1
9/94
7480
Rev 0
9/86

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
7481
7520
7550
7610
7740
7741
~ ~
7760
"
METH NO.
FINAL
UPDATE I
DATED
7/92
"
"

"
™* ~
~ ~
*" ~
7760A
7761
METH NO.
FINAL
UPDT. II
DATED
9/94
"" ~
*" ~
"™ ~"
"
"" ~
7741A
7742
"
"
METHOD TITLE
Molybdenum (Atomic
Absorption, Furnace
Technique)
Nickel (Atomic
Absorption, Direct
Aspiration)
Osmium (Atomic
Absorption, Direct
Aspiration)
Potassium (Atomic
Absorption, Direct
Aspiration)
Selenium (Atomic
Absorption, Furnace
Technique)
Selenium (Atomic
Absorption, Gaseous
Hydride)
Selenium (Atomic
Absorption,
Borohydride
Reduction)
Silver (Atomic
Absorption, Direct
Aspiration)
Silver (Atomic
Absorption, Furnace
Technique)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
7481
Rev 0
9/86
7520
Rev 0
9/86
7550
Rev 0
9/86
7610
Rev 0
9/86
7740
Rev 0
9/86
7741A
Rev 1
9/94
7742
Rev 0
9/94
7760A
Rev 1
7/92
7761
Rev 0
7/92
                                      10

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
7770
~ —
7840
7841
7870
7910
7911
7950
"
NETH NO.
FINAL
UPDATE I
DATED
7/92
*" ~
7780
— —
~ ~
~ ~
"
~* ~
"
7951
NETH NO.
FINAL
UPDT. II
DATED
9/94
~ **
~ ~
•" ~
~ ~
~ ~
"* ~
*" ~
"
"
METHOD TITLE
Sodium (Atomic
Absorption, Direct
Aspiration)
Strontium (Atomic
Absorption, Direct
Aspiration)
Thallium (Atomic
Absorption, Direct
Aspiration)
Thallium (Atomic
Absorption, Furnace
Technique)
Tin (Atomic
Absorption, Direct
Aspiration)
Vanadium (Atomic
Absorption, Direct
Aspiration)
Vanadium (Atomic
Absorption, Furnace
Technique)
Zinc (Atomic
Absorption, Direct
Aspiration)
Zinc (Atomic
Absorption, Furnace
Technique)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
7770
Rev 0
9/86
7780
Rev 0
7/92
7840
Rev 0
9/86
7841
Rev 0
9/86
7870
Rev 0
9/86
7910
Rev 0
9/86
7911
Rev 0
9/86
7950
Rev 0
9/86
7951
Rev 0
7/92
                                      11

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
8000
8010

8015
8020

8030
— —
NETH NO.
FINAL
UPDATE I
DATED
7/92
8000A
801 OA
8011
80 ISA
"* ~
8021
8030A
"
NETH NO.
FINAL
UPDT. II
DATED
9/94
"
8010B

*" *"
8020A
8021A
™ «•
8031
METHOD TITLE
Gas Chromatography
Halogenated Volatile
Organics by Gas
Chromatography
1,2-Dibromoethane
and l,2-Dibromo-3-
chloropropane by
Microextraction and
Gas Chromatography
Nonhalogenated
Volatile Organics by
Gas Chromatography
Aromatic Volatile
Organics by Gas
Chromatography
Halogenated
Volatiles by Gas
Chromatography Using
Photoionization and
Electrolytic
Conductivity
Detectors in Series:
Capillary Column
Technique
Acrolein and
Acrylonitrile by Gas
Chromatography
Acrylonitrile by Gas
Chromatography
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
CURRENT
PROMUL-
GATED
METHOD
8000A
Rev 1
7/92
8010B
Rev 2 '
9/94
8011
Rev 0
7/92
80 ISA
Rev 1
7/92
8020A
Rev 1
9/94
8021A
Rev 1
9/94
8030A
Rev 1
7/92
8031
Rev 0
9/94
                                      12

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
— —
8040
8060

"
8080

8090
HETH NO.
FINAL
UPDATE I
DATED
7/92
*. _
8040A
lm ~

8070



NETH NO.
FINAL
UPDT. II
DATED
9/94
8032
"™ ~
~ ~
8061
"
8080A
8081

NETHOD TITLE
Acryl amide by Gas
Chromatography
Phenols by Gas
Chromatography
Phthalate Esters
Phthalate Esters by
Capillary Gas
Chromatography with
Electron Capture
Detection (GC/ECD)
Nitrosamines by Gas
Chromatography
Organochlorine Pes-
ticides and
Polychlorinated
Biphenyls by Gas
Chromatography
Organochlorine
Pesticides and PCBs
as Aroclors by Gas
Chromatography:
Capillary Column
Technique
Nitroaromatics and
Cyclic Ketones
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
CURRENT
PROMUL-
GATED
METHOD
8032
Rev 0
9/94
8040A
Rev 1
7/92
8060
Rev 0
9/86
8061
Rev 0
9/94
8070
Rev 0
7/92
8080A
Rev 1
9/94
8081
Rev 0
9/94
8090
Rev 0
9/86
                                      13

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
HETH NO.
THIRD ED
DATED
9/86
8100
— _•
8120

8140

8150

METH NO.
FINAL
UPDATE I
DATED
7/92
~ ~
8110
_ _

_ _
8141
8150A

METH NO.
FINAL
UPDT. II
DATED
9/94
~ ~
~ ~
8120A
8121
"
8141A
8150B
8151
METHOD TITLE
Polynuclear Aromatic
Hydrocarbons
Haloethers by Gas
Chromatography
Chlorinated
Hydrocarbons by Gas
Chromatography
Chlorinated
Hydrocarbons by Gas
Chromatography:
Capillary Column
Technique
Organophosphorus
Pesticides
Organophosphorus
Compounds by Gas
Chromatography:
Capillary Column
Technique
Chlorinated
Herbicides by Gas
Chromatography
Chlorinated
Herbicides by GC
Using Methyl ati on or
Pentafluorobenzyl-
ation Derivati-
zation: Capillary
Column Technique
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
CURRENT
PROMUL-
GATED
METHOD
8100
Rev 0
9/86
8110
Rev 0
7/92
8120A
Rev 1
9/94
8121
Rev 0
9/94
8140
Rev 0
9/86
8141A
Rev 1
9/94
8150B
Rev 2
9/94
8151
Rev 0
9/94
                                      14

-------
SW-846 METHOD STATUS TABLE (9/94),  CONTINUED
METH NO.
THIRD ED
DATED
9/86
8240
8250

8270

8280
METH NO.
FINAL
UPDATE I
DATED
7/92
8240A

8260
8270A


METH NO.
FINAL
UPDT. II
DATED
9/94
8240B
8250A
8260A
8270B
8275

METHOD TITLE
Volatile Organic
Compounds by Gas
Chromatography/Mass
Spectrometry (GC/MS)
Semivolatile Organic
Compounds
by Gas
Chromatography/Mass
Spectrometry (GC/MS)
Volatile Organic
Compounds by Gas
Chromatography/Mass
Spectrometry
(GC/MS): Capillary
Column Technique
Semivolatile Organic
Compounds by Gas
Chromatography/Mass
Spectrometry
(GC/MS): Capillary
Column Technique
Thermal
Chromatography/Mass
Spectrometry (TC/MS)
for Screening
Semivolatile Organic
Compounds
The Analysis of
Polychlorinated
Dibenzo-p-Dioxins
and Polychlorinated
Dibenzofurans
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec 4.4
Vol IB
Chap 4
Sec
4.3.2
CURRENT
PROMUL-
GATED
METHOD
8240B
Rev 2
9/94
8250A
Rev 1
9/94
8260A
Rev 1
9/94
8270B
Rev 2
9/94
8275
Rev 0
9/94
8280
Rev 0
9/86
                                      15

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86

8310


"
METH NO.
FINAL
UPDATE I
DATED
7/92

_ •»


"
METH NO.
FINAL
UPDT. II
DATED
9/94
8290
"
8315
8316
8318
METHOD TITLE
Polychlorinated
Dibenzodioxins
(PCDDs) and
Polychlorlnated
Dibenzofurans
(PCDFs) by High-
Resolution Gas
Chromatography/High-
Resolution Mass
Spectrometry
(HRGC/HRMS)
Polynuclear Aromatic
Hydrocarbons
Determination of
Carbonyl Compounds
by High Performance
Liquid
Chromatography
(HPLC)
Aery 1 amide,
Acrylonitrile and
Acrolein by High
Performance Liquid
Chromatography
(HPLC)
N-Methyl carbamates
by High Performance
Liquid Chroma-
tography (HPLC)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
CURRENT
PROMUL-
GATED
METHOD
8290
Rev 0
9/94
8310
Rev 0
9/86
8315
Rev 0
9/94
8316
Rev 0
9/94
8318
Rev 0
9/94
                                                                                   €
                                      16

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86




9010
9012
NETH NO.
FINAL
UPDATE I
DATED
7/92




9010A
"
NETH NO.
FINAL
UPDT. II
DATED
9/94
8321
8330
8331
8410
~ "
"
NETHOD TITLE
Solvent Extractable
Non-Volatile
Compounds by High
Performance Liquid
Chromatography/Ther-
mospray/Mass
Spectrometry
(HPLC/TSP/MS) or
Ultraviolet (UV)
Detection
Nitroaromatics and
Nitramines by High
Performance Liquid
Chromatography
(HPLC)
Tetrazene by Reverse
Phase High
Performance Liquid
Chromatography
(HPLC)
Gas Chroma-
tography/Fourier
Transform Infrared
(GC/FT-IR) Spec-
trometry for
Semi volatile
Organics: Capillary
Column
Total and Amenable
Cyanide
(Colorimetric,
Manual)
Total and Amenable
Cyanide
(Colorimetric,
Automated UV)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.4
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
8321
Rev 0
9/94
8330
Rev 0
9/94
8331
Rev 0
9/94
8410
Rev 0
9/94
9010A
Rev 1
7/92
9012
Rev 0
9/86
                                      17

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
HETH NO.
THIRD ED
DATED
9/86
— —
9020
~ *~
9022
9030

9035
9036
9038
HETH NO.
FINAL
UPDATE I
DATED
7/92
9013
9020A
9021
-, _
9030A
9031
"

"
HETH NO.
FINAL
UPDT. II
DATED
9/94
— *•
9020B
"
~ "*
_ _
"
"


METHOD TITLE
Cyanide Extraction
Procedure for Solids
and Oils
Total Organic
Hal ides (TOX)
Purgeable Organic
Hal ides (POX)
Total Organic
Hal ides (TOX) by
Neutron Activation
Analysis
Acid-Soluble and
Acid-Insoluble
Sulfides
Extractable Sulfides
Sulfate
(Colorimetric,
Automated,
Chloranilate)
Sulfate
(Colorimetric,
Automated,
Methyl thymol Blue,
AA II)
Sulfate
(Turbidimetric)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
9013
Rev 0
7/92
9020B
Rev 2
9/94
9021
Rev 0
7/92
9022
Rev 0
9/86
9030A
Rev 1
7/92
9031
Rev 0
7/92
9035
Rev 0
9/86
9036
Rev 0
9/86
9038
Rev 0
9/86
                                      18

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9040
9041
9045
9050
"
9060
9065
9066
9067
NETH NO.
FINAL
UPDATE I
DATED
7/92
~ ~
9041A
9045A
~ ~
~" "*
•™ ~
*"* ™
"
"
NETH NO.
FINAL
UPDT. II
DATED
9/94
9040A
~ —
9045B
"
9056
~ ~
"

"
METHOD TITLE
pH Electrometric
Measurement
pH Paper Method
Soil and Waste pH
Specific Conductance
Determination of
Inorganic An ions by
Ion Chromatography
Total Organic Carbon
Phenol ics
(Spectrophotometri c ,
Manual 4-AAP with
Distillation)
Phenol ics
(Colorimetric,
Automated 4-AAP with
Distillation)
Phenol ics
(Spectrophotometri c,
MBTH with
Distillation)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
9040A
Rev 1
9/94
9041A
Rev 1
7/92
9045B
Rev 2
9/94
9050
Rev 0
9/86
9056
Rev 0
9/94
9060
Rev 0
9/86
9065
Rev 0
9/86
9066
Rev 0
9/86
9067
Rev 0
9/86
                                      19

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9070
9071



9080
9081
NETH NO.
FINAL
UPDATE I
DATED
7/92





— —
"
METH NO.
FINAL
UPDT. II
DATED
9/94

9071A
9075
9076
9077
"" ~
"
METHOD TITLE
Total Recoverable
Oil & Grease
(Gravimetric,
Separatory Funnel
Extraction)
Oil and Grease
Extraction Method
for Sludge and
Sediment
Samples
Test Method for
Total Chlorine in
New and Used
Petroleum Products
by X-Ray
Fluorescence
Spectrometry (XRF)
Test Method for
Total Chlorine in
New and Used
Petroleum Products
by Oxidative
Combustion and
Microcoulometry
Test Methods for
Total Chlorine in
New and Used
Petroleum Products
(Field Test Kit
Methods)
Cation- Exchange
Capacity of Soils
(Ammonium Acetate)
Cation-Exchange
Capacity of Soils
(Sodium Acetate)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 6
Vol 1C
Chap 6
CURRENT
PROMUL-
GATED
METHOD
9070
Rev 0
9/86
9071A
Rev 1
9/94
9075
Rev 0
9/94
9076
Rev 0
9/94
9077
Rev 0
9/94
9080
Rev 0
9/86
9081
Rev 0
9/86
                                      20

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9090
9095
™* ~
9100
9131
9132
9200
9250
9251
NETH NO.
FINAL
UPDATE I
DATED
7/92
9090A
*™ ~
_ _

— _
_ _

-» ~
"
METH NO.
FINAL
UPDT. II
DATED
9/94
~ ~
™* ••
9096

~ ~
~ ~
"
"

METHOD TITLE
Compatibility Test
for Wastes and
Membrane Liners
Paint Filter Liquids
Test
Liquid Release Test
(LRT) Procedure
Saturated Hydraulic
Conductivity,
Saturated Leachate
Conductivity, and
Intrinsic
Permeability
Total Col i form:
Multiple Tube
Fermentation
Technique
Total Col i form:
Membrane Filter
Technique
Nitrate
Chloride
(Colorimetric,
Automated
Ferri cyanide AAI)
Chloride
(Colorimetric,
Automated
Ferri cyanide AAI I)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
9090A
Rev 1
7/92
9095
Rev 0
9/86
9096
Rev 0
9/94
9100
Rev 0
9/86
9131
Rev 0
9/86
9132
Rev 0
9/86
9200
Rev 0
9/86
9250
Rev 0
9/86
9251
Rev 0
9/86
                                      21

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9252
_ _
9310
9315
9320
HCN Test
Method
H2S Test
Method
METH NO.
FINAL
UPDATE I
DATED
7/92
~ ~
™ ~
** ~
~ ~
_ _
HCN Test
Method
H2S Test
Method
METH NO.
FINAL
UPDT. II
DATED
9/94
9252A
9253
~ —
** ~
~" ~
HCN Test
Method
H2S Test
Method
NETHOD TITLE
Chloride
(Titrimetric,
Mercuric Nitrate)
Chloride
(Titrimetric, Silver
Nitrate)
Gross Alpha and
Gross Beta
Alpha-Emitting
Radium Isotopes
Radium- 228
Test Method to
Determine Hydrogen
Cyanide Released
from Wastes
Test Method to
Determine Hydrogen
Sulfide Released
from Wastes
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 5
Vol 1C
Chap 7
Sec 7.3
Vol 1C
Chap 7
Sec 7.3
CURRENT
PROMUL-
GATED
METHOD
9252A
Rev 1
9/94
9253
Rev 0
9/94
9310
Rev 0
9/86
9315
Rev 0
9/86
9320
Rev 0
9/86
Guidance
Method
Only
Guidance
Method
Only
€
                                      22

-------
                                        	                             a./

                                        VOLUME  ONE,


                                             SECTION A
|
                   U s  Environmental Protection Agency

                   Region 5, Library (PL-12J)
                   77 West Jackson Boulevard, 12UUlOor           „           --v   ,..f

                   fhiraso IL  60604-3590                       Revision       0
                   Chicago, IL  ou                               Date

-------
U.S. eiwlfonmwtil Pwtectkm Agency
                                                                                                        i
                     For rale by the Superintendent of Documents, tl.S. Government Printing Office Watblnirton, D.C. 20402

-------
                                  DISCLAIMER
      Mention  of  trade names  or  commercial   products  does  not  constitute
endorsement  or recommendation  for  use by  the U.S.  Environmental  Protection
Agency.

      SW-846 methods are designed to be used with equipment from any manufacturer
that results in suitable method performance (as  assessed by accuracy, precision,
detection limits and matrix compatibility).  In several  SW-846 methods, equipment
specifications  and  settings  are given for the  specific  instrument used during
method development,  or subsequently  approved  for use in  the  method.   These
references are made to provide the best possible guidance to laboratories using
this manual.  Equipment not specified  in the method may be used as long as the
laboratory  achieves  equivalent  or superior method performance.   If alternate
equipment is used, the laboratory must  follow the manufacturer's instructions for
their particular instrument.

      Since  many  types and  sizes  of  glassware and supplies are commercially
available, and  since  it is  possible to prepare reagents  and  standards in many
different ways, those specified in these methods may be replaced by any similar
types as  long  as this  substitution does not affect the overall  quality of the
analyses.
                                DISCLAIMER - 1                      Revision 0
                                                                     July 1992

-------
                                   ABSTRACT

Test Methods  for Evaluating Solid Waste,  Physical/Chemical  Methods   (SW-846)
provides test procedures and guidance which are recommended for use in conducting
the evaluations and measurements needed to comply with the Resource Conservation
and Recovery  Act  (RCRA),  Public Law 94-580, as  amended.    These methods  are
approved  by  the U.S.  Environmental  Protection  Agency for  obtaining data  to
satisfy the requirements of 40 CFR Parts 122 through 270 promulgated  under RCRA,
as amended.   This  manual presents the  state-of-the-art in routine  analytical
tested  adapted  for the  RCRA  program.    It contains  procedures  for field  and
laboratory quality control,  sampling,  determining  hazardous  constituents  in
wastes,  determining   the  hazardous   characteristics  of  wastes   (toxicity,
ignitability,  reactivity,  and  corrosivity),  and   for determining  physical
properties of wastes.   It  also  contains guidance on how to select  appropriate
methods.

      Several of the hazardous waste regulations  under Subtitle C of RCRA require
that  specific testing  methods   described  in SW-846  be employed for  certain
applications.  Refer to 40 Code of Federal  Regulations (CFR), Parts  260 through
270, for those specific requirements. Any reliable analytical method may be used
to meet other requirements under Subtitle  C  of RCRA.
                                  U.S. Environmental Protection Agency
                                  Region 5, Library (PL. 12j)
                                 ABSTRACT - 1                       Revision 2
                                                                September  1994

-------
                           TABLE   OF   CONTENTS
                                     VOLUME ONE

                                      SECTION A
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE
ACKNOWLEDGEMENTS
                    PART  I     METHODS  FOR ANALYTES AND  PROPERTIES

CHAPTER ONE -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References

CHAPTER TWO -- CHOOSING THE CORRECT PROCEDURE

      2.1   Purpose
      2.2   Required Information
      2.3   Implementing the Guidance
      2.4   Characteristics
      2.5   Ground Water
      2.6   References

CHAPTER THREE -- METALLIC ANALYTES

      3.1   Sampling Considerations
      3.2   Sample Preparation Methods

            Method 3005A:     Acid  Digestion of  Waters  for  Total  Recoverable  or
                              Dissolved Metals for Analysis by Flame Atomic Absorption
                              (FLAA) or Inductively Coupled Plasma (ICP) Spectroscopy
            Method 3010A:     Acid Digestion of Aqueous Samples and Extracts for Total
                              Metals for Analysis by Flame Atomic Absorption (FLAA) or
                              Inductively Coupled Plasma (ICP) Spectroscopy
            Method 3015:       Microwave Assisted Acid Digestion of Aqueous Samples and
                              Extracts
                                    CONTENTS - 1                          Revision 2
                                                                      September 1994

-------
      Method 3020A:
      Method 3040:
      Method 3050A:
      Method 3051:
           Acid Digestion of Aqueous Samples and Extracts for Total
           Metals  for   Analysis   by  Graphite   Furnace   Atomic
           Absorption (GFAA)  Spectroscopy
           Dissolution Procedure for Oils, Greases, or Waxes
           Acid Digestion of Sediments, Sludges, and Soils
           Microwave Assisted Acid Digestion of Sediments, Sludges,
           Soils, and Oils
3.3   Methods for Determination of Metals
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
6010A:
6020:
7000A:
7020:
7040:
7041:
7060A:
7061A:
7062:
7080A:
7081:
7090:
7091:
7130:
7131A:
7140:
7190:
7191:
7195:
7196A:
7197:
7198:
7200:
7201:
7210:
7211:
7380:
7381:
7420:
7421:
7430:
7450:
7460:
7461:
7470A:
7471A:
      Method 7480:
      Method 7481:
      Method 7520:
      Method 7550:
      Method 7610:
      Method 7740:
Inductively Coupled Plasma-Atomic Emission Spectroscopy
Inductively Coupled Plasma - Mass Spectrometry
Atomic Absorption Methods
Aluminum (AA, Direct Aspiration)
Antimony (AA, Direct Aspiration)
Antimony (AA, Furnace Technique)
Arsenic (AA, Furnace Technique)
Arsenic (AA, Gaseous Hydride)
Antimony and Arsenic (AA, Borohydride Reduction)
Barium (AA, Direct Aspiration)
Barium (AA, Furnace Technique)
Beryllium (AA, Direct Aspiration)
Beryllium (AA, Furnace Technique)
Cadmium (AA, Direct Aspiration)
Cadmium (AA, Furnace Technique)
Calcium (AA, Direct Aspiration)
Chromium (AA, Direct Aspiration)
Chromium (AA, Furnace Technique)
Chromium, Hexavalent (Coprecipitation)
Chromium, Hexavalent (Colorimetric)
Chromium, Hexavalent (Chelation/Extraction)
Chromium, Hexavalent (Differential Pulse Polarography)
Cobalt (AA, Direct Aspiration)
Cobalt (AA, Furnace Technique)
Copper (AA, Direct Aspiration)
Copper (AA, Furnace Technique)
Iron (AA, Direct Aspiration)
Iron (AA, Furnace Technique)
Lead (AA, Direct Aspiration)
Lead (AA, Furnace Technique)
Lithium (AA, Direct Aspiration)
Magnesium (AA, Direct Aspiration)
Manganese (AA, Direct Aspiration)
Manganese (AA, Furnace Technique)
Mercury in  Liquid Waste  (Manual Cold-Vapor Technique)
Mercury in  Solid or Semisolid Waste  (Manual Cold-Vapor
Technique)
Molybdenum  (AA, Direct Aspiration)
Molybdenum  (AA, Furnace  Technique)
Nickel (AA, Direct Aspiration)
Osmium (AA, Direct Aspiration)
Potassium (AA, Direct Aspiration)
Selenium (AA, Furnace Technique)
                              CONTENTS - 2
                                                       Revision 2
                                                   September  1994

-------
            Method 7741A:     Selenium (AA, Gaseous Hydride)
            Method 7742:      Selenium (AA, Borohydride Reduction)
            Method 7760A:     Silver (AA, Direct Aspiration)
            Method 7761:      Silver (AA, Furnace Technique)
            Method 7770:      Sodium (AA, Direct Aspiration)
            Method 7780:      Strontium (AA, Direct Aspiration)
            Method 7840:      Thallium (AA, Direct Aspiration)
            Method 7841:      Thallium (AA, Furnace Technique)
            Method 7870:      Tin (AA, Direct Aspiration)
            Method 7910:      Vanadium (AA, Direct Aspiration)
            Method 7911:      Vanadium (AA, Furnace Technique)
            Method 7950:      Zinc (AA, Direct Aspiration)
            Method 7951:      Zinc (AA, Furnace Technique)
APPENDIX -- COMPANY REFERENCES
        NOTE;  A  suffix of "A" in the method  number  indicates  revision one
        (the method  has been  revised once).  A suffix of  "B"  in the method
        number indicates revision two (the method has been revised twice). In
        order to  properly  document the method  used  for  analysis, the entire
        method number Including the suffix letter designation (e.g., A or B)
        must be Identified by the analyst.   A  method  reference  found within
        the  RCRA  regulations  and the text  of SW-846 methods  and  chapters
        refers to the latest promulgated revision of the method, even though
        the method number  does not include the appropriate letter suffix.
                                    CONTENTS - 3                          Revision 2
                                                                      September 1994

-------
                                     VOLUME ONE

                                      SECTION  B
i
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE
ACKNOWLEDGEMENTS

CHAPTER ONE. REPRINTED -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References

CHAPTER FOUR -- ORGANIC ANALYTES

      4.1   Sampling Considerations
      4.2   Sample Preparation Methods

            4.2.1       Extractions and Preparations

            Method 3500A:     Organic Extraction and Sample Preparation
            Method 3510B:     Separatory Funnel Liquid-Liquid Extraction
            Method 3520B:     Continuous Liquid-Liquid Extraction
            Method 3540B:     Soxhlet Extraction
            Method 3541:      Automated Soxhlet Extraction
            Method 3550A:     Ultrasonic Extraction
            Method 3580A:     Waste Dilution
            Method 5030A:     Purge-and-Trap
            Method 5040A:     Analysis  of  Sorbent Cartridges from  Volatile Organic
                              Sampling  Train   (VOST):     Gas   Chromatography/Mass
                              Spectrometry Technique
            Method 5041:      Protocol  for  Analysis  of  Sorbent  Cartridges  from
                              Volatile  Organic Sampling  Train  (VOST):    Wide-bore
                              Capillary Column Technique
            Method 5100:      Determination of the Volatile Organic Concentration of
                              Waste Samples
            Method 5110:      Determination of Organic Phase Vapor Pressure in Waste
                              Samples

            4.2.2       Cleanup

            Method 3600B:     Cleanup
            Method 3610A:     Alumina Column Cleanup
                                    CONTENTS - 4                          Revision 2
                                                                      September 1994

-------
      Method 3611A:
      Method
      Method
      Method
      Method
      Method
      Method
3620A:
3630B:
3640A:
3650A:
3660A:
3665:
Alumina    Column    Cleanup     and    Separation    of
Petroleum Wastes
Florisil Column Cleanup
Silica Gel Cleanup
Gel-Permeation Cleanup
Acid-Base Partition Cleanup
Sulfur Cleanup
Sulfuric Acid/Permanganate Cleanup
4.3   Determination of Organic Analytes
      4.3.1
     Gas Chromatographic Methods
      Method 8000A:
      Method 8010B:
      Method 8011:

      Method 8015A:
      Method 8020A:
      Method 8021A:
      Method
      Method
      Method
      Method
      Method
      Method
8030A:
8031:
8032:
8040A:
8060:
8061:
      Method 8070:
      Method 8080A:

      Method 8081:

      Method 8090:
      Method 8100:
      Method 8110:
      Method 8120A:
      Method 8121:

      Method 8140:
      Method 8141A:

      Method 81SOB:
      Method 8151:
Gas Chromatography
Halogenated Volatile Organics by Gas Chromatography
1,2-Dibromoethane  and l,2-Dibromo-3-chloropropane  by
Microextraction and Gas Chromatography
Nonhalogenated Volatile Organics by Gas Chromatography
Aromatic Volatile Organics by Gas Chromatography
Halogenated  Volatiles  by  Gas  Chromatography  Using
Photoionization and Electrolytic Conductivity Detectors
in Series: Capillary Column Technique
Acrolein and Acrylonitrile by Gas Chromatography
Acrylonitrile by Gas Chromatography
Acrylamide by Gas Chromatography
Phenols by Gas Chromatography
Phthalate Esters
Phthalate Esters by  Capillary  Gas  Chromatography with
Electron Capture Detection (GC/ECD)
Nitrosamines by Gas Chromatography
Organochlorine Pesticides and Polychlorinated Biphenyls
by Gas Chromatography
Organochlorine Pesticides and  PCBs  as  Aroclors by Gas
Chromatography:  Capillary Column Technique
Nitroaromatics and Cyclic Ketones
Polynuclear Aromatic Hydrocarbons
Haloethers by Gas Chromatography
Chlorinated Hydrocarbons by Gas Chromatography
Chlorinated   Hydrocarbons   by  Gas   Chromatography:
Capillary Column Technique
Organophosphorus Pesticides
Organophosphorus  Compounds  by   Gas  Chromatography:
Capillary Column Technique
Chlorinated Herbicides by Gas Chromatography
Chlorinated  Herbicides  by   GC Using  Methylation  or
Pentaf 1 uorobenzylation Deri vatization:  Capi 11 ary Column
Technique
                              CONTENTS - 5
                                                       Revision 2
                                                   September 1994

-------
4.3.2
Gas Chromatographic/Mass Spectroscopic Methods
Method 8240B:

Method 8250A:

Method 8260A:

Method 8270B:


Method 8280:

      Appendix A:
      Appendix B:

Method 8290:



      Appendix A:
      Volatile Organic Compounds by Gas  Chromatography/Mass
      Spectrometry (GC/MS)
      Semivolatile     Organic     Compounds    by    Gas
      Chromatography/Mass Spectrometry (GC/MS)
      Volatile Organic Compounds by Gas  Chromatography/Mass
      Spectrometry (GC/MS):  Capillary  Column  Technique
      Semivolatile     Organic     Compounds    by    Gas
      Chromatography/Mass  Spectrometry  (GC/MS):   Capillary
      Column Technique
      The Analysis of  Polychlorinated  Dibenzo-p-Dioxins
      Polychlorinated Dibenzofurans
            Signal-to-Noise Determination Methods
            Recommended Safety and  Handling  Procedures
            PCDDs/PCDFs
      Polychlorinated    Dibenzodioxins    (PCDDs)
      Polychlorinated Dibenzofurans (PCDFs) by High-Resolution
      Gas  Chromatography/High-Resolution  Mass  Spectrometry
      (HRGC/HRMS)
            Procedures   for   the   Collection,    Handling,
            Analysis, and Reporting of Wipe  Tests  Performed
            within the Laboratory
and
for

and
4.3.3

Method 8310:
Method 8315:
      Appendix A:

Method 8316:

Method 8318:

Method 8321:


Method 8330:

Method 8331:
High Performance Liquid Chromatographic Methods
      Polynuclear Aromatic Hydrocarbons
      Determination of Carbonyl Compounds by High Performance
      Liquid Chromatography (HPLC)
            Recrystallization of  2,4-Dinitrophenylhydrazine
            (DNPH)
      Acrylamide,   Acrylonitrile   and  Acrolein   by   High
      Performance Liquid Chromatography (HPLC)
      N-Methylcarbamates   by   High    Performance    Liquid
      Chromatography (HPLC)
      Solvent  Extractable Non-Volatile  Compounds  by  High
      Performance   Liquid   Chromatography/Thermospray/Mass
      Spectrometry (HPLC/TSP/MS) or Ultraviolet (UV) Detection
      Nitroaromatics and Nitramines by High Performance Liquid
      Chromatography (HPLC)
      Tetrazene  by Reverse  Phase High  Performance  Liquid
      Chromatography (HPLC)
4.3.4

Method 8410:
Fourier Transform Infrared Methods

      Gas Chromatography/Fourier Transform Infrared (GC/FT-IR)
      Spectrometry  for  Semivolatile  Organics:    Capillary
      Column
                        CONTENTS - 6
                                                  Revision 2
                                              September 1994

-------
      4.4   Miscellaneous Screening Methods

            Method 3810:      Headspace
            Method 3820:      Hexadecane  Extraction  and   Screening   of  Purgeable
                              Organics
            Method 4010:      Screening for Pentachlorophenol by Immunoassay
            Method 8275:      Thermal  Chromatography/Mass  Spectrometry  (TC/MS)  for
                              Screening Semivolatile Organic Compounds

APPENDIX -- COMPANY REFERENCES
        NOTE;  A  suffix of "A" in the method number  indicates  revision one
        (the method has been  revised once).   A suffix of  "B" in  the method
        number indicates revision two (the method has been revised twice). In
        order to properly  document the method used  for  analysis,  the entire
        method number Including the suffix letter designation (e.g., A or B)
        must be identified by  the analyst.   A method  reference  found within
        the  RCRA  regulations  and the text  of  SW-846 methods  and  chapters
        refers to the latest promulgated revision of the method, even though
        the method number does not include the appropriate letter suffix.
                                    CONTENTS -  7                          Revision 2
                                                                      September 1994

-------
                                     VOLUME ONE

                                      SECTION  C
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE

CHAPTER ONE, REPRINTED -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References

CHAPTER FIVE -- MISCELLANEOUS TEST METHODS
            Method
            Nethod
            Method
            Method
            Method
            Method
            Method
5050:
9010A:
9012:
9013:
9020B:
9021:
9022:
            Method 9030A:
            Method 9031:
            Nethod 9035:
            Method 9036:

            Method 9038:
            Method 9056:
            Method 9060:
            Method 9065:

            Method 9066:

            Method 9067:
            Method 9070:

            Method 9071A:

            Method 9075:

            Method 9076:
Bomb Preparation Method for Solid Waste
Total and Amenable Cyanide (Colorimetric, Manual)
Total and Amenable Cyanide  (Colorimetric, Automated UV)
Cyanide Extraction Procedure for Solids and Oils
Total Organic Hal ides (TOX)
Purgeable Organic Hal ides (POX)
Total  Organic  Hal ides   (TOX)  by  Neutron  Activation
Analysis
Acid-Soluble and Acid-Insoluble Sulfides
Extractable Sulfides
Sulfate (Colorimetric, Automated, Chloranilate)
Sulfate (Colorimetric, Automated, Methylthymol Blue, AA
II)
Sulfate (Turbidimetric)
Determination of Inorganic Anions by Ion Chromatography
Total Organic Carbon
Phenolics   (Spectrophotometric,   Manual   4-AAP  with
Distillation)
Phenolics    (Colorimetric,   Automated   4-AAP   with
Distillation)
Phenolics (Spectrophotometric,  MBTH with Distillation)
Total Recoverable Oil & Grease  (Gravimetric, Separatory
Funnel Extraction)
Oil and Grease Extraction Method for Sludge  and Sediment
Samples
Test Method for  Total Chlorine in New and  Used Petroleum
Products by X-Ray Fluorescence Spectrometry (XRF)
Test Method for  Total Chlorine in New and  Used Petroleum
Products by Oxidative Combustion and Microcoulometry
i
                                    CONTENTS  - 8
                                                       Revision 2
                                                   September 1994

-------
            Method 9077:

                  Method A:
                  Method B:

                  Method C:
            Method 9131:
            Method 9132:
            Method 9200:
            Method 9250:
            Method 9251:
            Method 9252A:
            Method 9253:
            Method 9320:

CHAPTER SIX -- PROPERTIES

            Method 1312:
            Method 1320:
            Method 1330A:
            Method 9040A:
            Method 9041A:
            Method 9045B:
            Method 9050:
            Method 9080:
            Method 9081:
            Method 9090A:
            Method 9095:
            Method 9096:
                  Appendix A:
            Method 9100:

            Method 9310:
            Method 9315:
Test  Methods  for  Total  Chlorine   in  New  and  Used
Petroleum Products (Field Test Kit Methods)
      Fixed End Point Test Kit Method
      Reverse Titration Quantitative End Point Test Kit
      Method
Direct Titration Quantitative End  Point Test Kit Method
Total Coliform:  Multiple Tube Fermentation Technique
Total Coliform:  Membrane Filter Technique
Nitrate
Chloride (Colorimetric, Automated Ferricyanide AAI)
Chloride (Colorimetric, Automated Ferricyanide AAII)
Chloride (Titrimetric, Mercuric Nitrate)
Chloride (Titrimetric, Silver Nitrate)
Radium-228
Synthetic Precipitation Leaching Procedure
Multiple Extraction Procedure
Extraction Procedure for Oily Wastes
pH Electrometric Measurement
pH Paper Method
Soil and Waste pH
Specific Conductance
Cation-Exchange Capacity of Soils (Ammonium Acetate)
Cation-Exchange Capacity of Soils (Sodium Acetate)
Compatibility Test for Wastes and Membrane Liners
Paint Filter Liquids Test
Liquid Release Test (LRT) Procedure
      LRT Pre-Test
Saturated Hydraulic  Conductivity,  Saturated  Leachate
Conductivity, and Intrinsic Permeability
Gross Alpha and Gross Beta
Alpha-Emitting Radium Isotopes
                              PART II    CHARACTERISTICS

CHAPTER SEVEN -- INTRODUCTION AND REGULATORY DEFINITIONS

      7.1   Ignitability
      7.2   Corrosivity
      7.3   Reactivity

            Test Method to Determine Hydrogen Cyanide Released from Wastes
            Test Method to Determine Hydrogen Sulfide Released from Wastes

      7.4   Toxicity Characteristic Leaching Procedure
                                    CONTENTS - 9
                                            Revision 2
                                        September 1994

-------
CHAPTER EIGHT -- METHODS FOR DETERMINING CHARACTERISTICS

      8.1   Ignitability

            Method 1010:      Pensky-Martens   Closed-Cup   Method   for   Determining
                              Ignitability
            Method 1020A:     Setaflash Closed-Cup Method for Determining Ignitability

      8.2   Corrosivity

            Method 1110:      Corrosivity Toward Steel

      8.3   Reactivity
      8.4   Toxicity

            Method 1310A:     Extraction  Procedure   (EP)  Toxicity  Test  Method  and
                              Structural Integrity Test
            Method 1311:      Toxicity Characteristic Leaching Procedure

APPENDIX -- COMPANY REFERENCES
        NOTE;   A suffix of "A" in the method  number  indicates  revision one
        (the method  has been  revised once).  A suffix of  "B"  in  the method
        number  indicates revision two (the method  has been revised twice). In
        order to  properly  document the method  used for  analysis,  the entire
        method  number Including the suffix letter designation (e.g., A or B)
        must be Identified by the analyst.   A  method  reference found within
        the  RCRA regulations  and the text  of SW-846 methods  and  chapters
        refers  to the latest  promulgated revision of the method, even though
        the method number  does not include the appropriate letter suffix.
                                    CONTENTS - 10                          Revision 2
                                                                      September  1994

-------
                                     VOLUME   TWO
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE
CHAPTER ONE. REPRINTED -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References
                                 PART  III    SAMPLING

CHAPTER NINE -- SAMPLING PLAN

      9.1   Design and Development
      9.2   Implementation

CHAPTER TEN -- SAMPLING METHODS

            Method 0010:      Modified Method 5 Sampling Train
                  Appendix A:       Preparation of XAD-2 Sorbent Resin
                  Appendix B:       Total Chromatographable Organic Material Analysis
            Method 0020:      Source Assessment Sampling System (SASS)
            Method 0030:      Volatile Organic Sampling Train
                                PART IV   MONITORING

CHAPTER ELEVEN -- GROUND WATER MONITORING

      11.1  Background and Objectives
      11.2  Relationship to the Regulations and to Other Documents
      11.3  Revisions and Additions
      11.4  Acceptable Designs and Practices
      11.5  Unacceptable Designs and Practices


CHAPTER TWELVE -- LAND TREATMENT MONITORING

      12.1  Background
      12.2  Treatment Zone
      12.3  Regulatory Definition
                                    CONTENTS  -  11                          Revision 2
                                                                      September 1994

-------
      12.4  Monitoring and Sampling Strategy
      12.5  Analysis
      12.6  References and Bibliography

CHAPTER THIRTEEN - INCINERATION

      13.1  Introduction
      13.2  Regulatory Definition
      13.3  Waste Characterization Strategy
      13.4  Stack-Gas Effluent Characterization Strategy
      13.5  Additional Effluent Characterization Strategy
      13.6  Selection of Specific Sampling and Analysis Methods
      13.7  References

APPENDIX -- COMPANY REFERENCES
        NOTE;  A  suffix of "A" in the method number  indicates  revision one
        (the method  has been  revised once).   A suffix of  "B"  in  the method
        number indicates revision two (the method has been revised twice). In
        order to  properly  document the method used  for  analysis,  the entire
        method number Including the suffix letter designation (e.g., A or B)
        must be Identified by the analyst.   A method  reference found within
        the  RCRA  regulations  and the text  of  SW-846 methods  and  chapters
        refers to the latest promulgated revision of the method, even though
        the method number  does not include the appropriate letter suffix.
                                    CONTENTS -  12                          Revision 2
                                                                      September 1994

-------
                            PREFACE AND OVERVIEW
PURPOSE OF THE MANUAL
     Test Methods for Evaluating Solid Waste (SW-846)  1s Intended to provide a
unified, up-to-date source of Information  on sampling and analysis related to
compliance with RCRA regulations.   It  brings together Into one reference all
sampling and testing methodology approved by the Office of Solid Waste for use
1n Implementing the RCRA regulatory  program.  The manual provides methodology
for collecting and testing representative samples of waste and other materials
to be monitored.  Aspects  of  sampling  and testing covered in SW-846 Include
quality control, sampling  plan  development  and  implementation, analysis of
Inorganic and  organic  constituents,  the  estimation  of  Intrinsic physical
properties, and the appraisal of waste characteristics.

     The procedures described 1n this manual are meant to be comprehensive and
detailed, coupled  with  the  realization  that  the  problems  encountered in
sampling and analytical situations  require  a  certain amount of flexibility.
The solutions to these problems will  depend, 1n part, on the skill, training,
and experience of the analyst.    For  some  situations, it is possible to use
this manual  1n  rote  fashion.    In  other  situations,  it  will  require a
combination of technical abilities, using  the  manual as guidance rather than
in a step-by-step, word-by-word fashion.    Although this puts an extra burden
on the  user,  1t  is  unavoidable  because  of  the  variety  of sampling and
analytical conditions found with hazardous wastes.
ORGANIZATION AND FORMAT
     This manual  is divided  Into two  volumes.  Volume I focuses on laboratory
activities and  is divided  for  convenience  into  three  sections.  Volume IA
deals  with  quality  control,  selection  of  appropriate  test  methods, and
analytical methods for metallic species.    Volume  IB consists of methods for
organic  analytes.    Volume 1C  includes  a  variety  of  test  methods  for
miscellaneous   analytes  and properties  for  use  in  evaluating  the  waste
characteristics.  Volume II  deals with sample acquisition and Includes quality
control, sampling plan design and  implementation, and field sampling methods.
Included for the  convenience of  sampling  personnel  are discusssions of the
ground water, land treatment, and incineration monitoring regulations.

     Volume I begins with an overview  of the quality control procedures to be
imposed upon the  sampling and  analytical methods. The quality control chapter
(Chapter One) and the  methods  chapters  are  interdependent.  The analytical
procedures cannot be  used   without  a  thorough  understanding of the quality
control requirements and the means to  implement them.  This understanding can
be  achieved only  be reviewing Chapter One and the analytical methods together.
It  1s expected  that  Individual  laboratories,  using  SW-846 as the reference
                                  PREFACE -  1
                                                          Revision      0
                                                          Date  September 1986

-------
source, will  select  appropriate  methods  and  develop  a standard operating
procedure (SOP) to be followed by  the laboratory.  The SOP should Incorporate
the pertinent Information from this  manual  adopted to the specific needs and
circumstances of the Individual laboratory as  well  as to the materials to be
evaluated.

     The method  selection  chapter  (Chapter  Two)  presents  a comprehensive
discussion of the application  of  these  methods  to  various matrices 1n the
determination of groups of analytes or specific analytes.  It aids the chemist
In constructing the correct  analytical  method  from  the array of procedures
which may  cover  the  matrlx/analyte/concentratlon  combination of Interests.
The  section  discusses  the  objective   of   the  testing  program  and  Us
relationship to the choice of an analytical method.  Flow charts are presented
along with  tables  to  guide  1n  the  selection  of  the  correct analytical
procedures  to form the appropriate method.

     The  analytical methods are  separated  Into distinct procedures describing
specific,   independent  analytical  operations.    These  include  extraction,
digestion,  cleanup, and  determination.     This   format  allows linking of the
various steps  in the analysis  according   to:   the  type of sample  (e.g., water,
soil,  sludge,  still bottom); analytes(s)   of  Interest; needed  sensitivity; and
available analytical instrumentation.     The  chapters describing Miscellaneous
Test Methods and   Properties,  however,   give complete  methods which  are not
amenable  to such  segmentation  to form discrete procedures.

     The  Introductory material   at  the   beginning of each  section  containing
analytical   procedures    presents    Information   on    sample   handling and
preservation,  safety, and  sample preparation.

     Part  II   of  Volume    I    (Chapters   Seven and   Eight)  describes the
characteristics of a waste.    Sections   following the regulatory  descriptions
contain  the methods  used to   determine  if  the   waste  1s  hazardous  because  1t
exhibits  a  particular characteristic.

     Volume II gives background   information  on  statistical  and nonstatistlcal
aspects   of  sampling.     It   also   presents practical   sampling  techniques
appropriate for situations presenting a  variety  of physical  conditions.

     A discussion  of   the  regulatory  requirements   with   respect  to several
monitoring  categories  is  also  given   1n  this  volume.    These include ground
water  monitoring,  land   treatment,   and   incineration.     The purpose of this
guidance is to orient the  user to  the objective  of the analysis,  and to assist
in developing  data quality objectives, sampling  plans,  and laboratory SOP's.

     Significant  interferences,  or  other  problems,   may  be encountered with
certain  samples.   In these situations,  the  analyst  is  advised to contact the
Chief, Methods Section  (WH-562B) Technical  Assessment Branch, Office of Solid
Waste, US EPA,  Washington,   DC    20460  (202-382-4761)   for assistance.  The
manual 1s intended to  serve   all  those   with  a need to evaluate solid waste.
 Your comments, corrections,  suggestions, and questions concerning any material
 contained in,  or omitted  from,   this  manual  will  be gratefully appreciated.
 Please direct your comments  to the above address.


                                  PREFACE - 2
                                                          Revision      0
                                                          Date  September 1986

-------
   CHAPTER ONE
TABLE OF CONTENTS
Secti
1.0
2.0















3.0






















on
INTRODUCTION 	
QA PROJECT PLAN 	
2.1 DATA QUALITY OBJECTIVES 	
2.2 PROJECT OBJECTIVES 	
2.3 SAMPLE COLLECTION 	
2.4 ANALYSIS AND TESTING 	
2.5 QUALITY CONTROL 	
2.6 PROJECT DOCUMENTATION 	
2.7 ORGANIZATION PERFORMING FIELD OR LABORATORY
OPERATIONS 	
2.7.1 Performance Evaluation 	
2.7.2 Internal Assessment by QA Function 	
2.7.3 External Assessment 	
2.7.4 On-Site Evaluation 	
2.7.4.1 Field Activities 	
2.7.4.2 Laboratory Activities 	
2.7.5 QA Reports 	
FIELD OPERATIONS 	
3.1 FIELD LOGISTICS 	
3.2 EQUIPMENT/INSTRUMENTATION 	
3.3 OPERATING PROCEDURES 	
3.3.1 Sample Management 	
3.3.2 Reagent/Standard Preparation 	
3.3.3 Decontamination 	
3.3.4 Sample Collection 	
3.3.5 Field Measurements 	
3.3.6 Equipment Calibration And Maintenance . . . .
3.3.7 Corrective Action 	
3.3.8 Data Reduction and Validation 	
3.3.9 Reporting 	
3.3.10 Records Management 	
3.3.11 Waste Disposal 	
3.4 FIELD QA AND QC REQUIREMENTS 	
3.4.1 Control Samples 	
3.4.2 Acceptance Criteria 	
3.4.3 Deviations 	
3.4.4 Corrective Action 	
3.4.5 Data Handling 	
3.5 QUALITY ASSURANCE REVIEW 	
3.6 FIELD RECORDS 	
Paqe
.... 1
.... 1
.... 2
.... 2
.... 3
.... 3
.... 3
.... 3

.... 4
.... 5
.... 5
.... 5
.... 5
.... 5
.... 6
.... 7
.... 8
.... 8
.... 9
.... 9
.... 9
.... 9
.... 9
.... 10
.... 10
.... 10
.... 10
.... 11
.... 11
.... 11
.... 11
.... 11
.... 11
.... 12
.... 12
.... 12
.... 12
.... 13
.... 13
     ONE - i                           Revision  1
                                      July  1992

-------
TABLE OF CONTENTS
   (continued)
Section
4.0























5.0
6.0
INDEX
LABORATORY OPERATIONS 	
4.1 FACILITIES 	
4.2 EQUIPMENT/INSTRUMENTATION 	
4.3 OPERATING PROCEDURES 	
4.3.1 Sample Management 	
4.3.2 Reagent/Standard Preparation 	
4.3.3 General Laboratory Techniques 	
4.3.4 Test Methods 	
4.3.5 Equipment Calibration and Maintenance . . . .
4.3.6 QC 	
4.3.7 Corrective Action 	
4.3.8 Data Reduction and Validation 	
4.3.9 Reporting 	
4.3.10 Records Management 	
4.3.11 Waste Disposal 	
4.4 LABORATORY QA AND QC PROCEDURES 	
4.4.1 Method Proficiency 	
4.4.2 Control Limits 	
4.4.3 Laboratory Control Procedures 	
4.4.4 Deviations 	
4.4.5 Corrective Action 	
4.4.6 Data Handling 	
4.5 QUALITY ASSURANCE REVIEW 	
4.6 LABORATORY RECORDS 	
DEFINITIONS 	
REFERENCES 	

Page
.... 14
.... 14
.... 15
.... 15
.... 16
.... 16
.... 16
.... 16
.... 17
.... 17
.... 17
.... 18
.... 18
.... 18
.... 18
.... 18
.... 18
.... 19
.... 19
.... 20
... 20
.... 20
.... 21
.... 21
.... 23
.... 29
.... 30
     ONE - ii                          Revision 1
                                       July 1992

-------
                                  CHAPTER ONE
                                QUALITY  CONTROL
1.0  INTRODUCTION
      It  is  the goal  of the U.S.  Environmental  Protection Agency's  (EPA's)
quality assurance (QA)  program to  ensure that all data be scientifically valid,
defensible,  and of  known precision  and  accuracy.   The  data  should be  of
sufficient known quality  to withstand scientific and  legal challenge relative to
the use for which the data are obtained.  The QA program is management's tool for
achieving this goal.

      For RCRA analyses,  the recommended minimum requirements for a QA program
and the associated  quality control (QC) procedures  are  provided in this chapter.

      The data acquired from QC procedures are used to estimate the quality of
analytical data, to  determine the  need for corrective  action  in  response to
identified  deficiencies, and  to  interpret results  after  corrective  action
procedures are implemented.   Method-specific QC procedures are incorporated in
the individual methods since they are not applied universally.

      A total program to  generate data of acceptable  quality should include both
a QA component, which encompasses the management procedures and controls, as well
as an  operational  day-to-day QC component.   This chapter  defines  fundamental
elements of such a data collection program.   Data collection efforts involve:

      1.    design of  a  project plan  to  achieve  the data quality  objectives
            (DQOs);

      2.    implementation of the project plan; and

      3.    assessment of the data to determine if the DQOs are met.

The project plan may  be a sampling and analysis plan  or a waste analysis plan if
it covers  the QA/QC  goals of the Chapter,  or  it  may be  a  Quality Assurance
Project Plan as described later in this chapter.

      This chapter identifies the minimal  QC components  that should be used in
the performance  of sampling  and analyses,  including  the  QC  information  which
should be documented.  Guidance is provided to construct QA programs for field
and laboratory work conducted in support of the RCRA program.


2.0  QA PROJECT PLAN

      It is recommended that all  projects which generate environment-related data
in support  of RCRA  have a   QA  Project  Plan (QAPjP)  or  equivalent.   In  some
instances,  a sampling  and   analysis  plan  or  a  waste  analysis  plan may  be
equivalent if it covers  all  of  the QA/QC goals outlined  in  this  chapter.   In
addition,  a  separate  QAPjP  need  not  be  prepared  for  routine  analyses  or
activities where  the  procedures  to be followed  are  described in  a Standard

                                    ONE  -  1                          Revision 1
                                                                     July 1992

-------
Operating Procedures manual or similar document and  include  the  elements  of a
QAPjP.  These documents should  be  available and referenced in the documentation
and/or records for the  analysis activities.   The  term  "QAPjP"  in this chapter
refers to any of these QA/QC documents.

      The QAPjP should detail  the  QA/QC goals  and protocols for a specific data
collection activity.   The QAPjP  sets forth  a plan for  sampling  and analysis
activities that will  generate data of a quality commensurate with their intended
use.   QAPjP  elements  should  include  a  description  of  the  project  and  its
objectives; a statement of the  DQOs  of the project; identification of those in-
volved  in the  data  collection  and their  responsibilities  and  authorities;
reference  to  (or inclusion  of)  the specific  sample collection  and analysis
procedures that will  be followed for all  aspects of the project; enumeration of
QC procedures  to  be  followed;  and  descriptions of all  project documentation.
Additional elements  should  be  included  in the QAPjP if  needed  to address all
quality related  aspects of the data  collection  project.  Elements  should be
omitted only when  they are inappropriate  for the project or when  absence of those
elements  will  not affect the  quality  of data obtained  for the  project  (see
reference 1).

      The role and importance  of  DQOs and project documentation  are discussed
below in Sections 2.1 through 2.6.  Management  and organization play a critical
role in determining the effectiveness of a QA/QC  program and ensuring that all
required  procedures  are followed.  Section 2.7 discusses the   elements of an
organization's QA program that have been found to ensure an effective program.
Field operations and  laboratory operations (along with applicable QC procedures)
are discussed  in Sections 3 and 4,  respectively.


2.1  DATA QUALITY OBJECTIVES

      Data quality objectives (DQOs) for the data collection activity describe
the overall level of uncertainty that a decision-maker is willing to accept in
results derived from  environmental data.   This  uncertainty is used to specify the
quality of  the measurement  data required, usually in  terms  of objectives for
precision, bias, representativeness, comparability and completeness.  The DQOs
should be defined prior to the  initiation of the field and laboratory  work.  The
field and  laboratory organizations  performing  the  work should  be aware of the
DQOs so that their personnel may make informed decisions during the course of the
project to attain those DQOs.   More detailed  information on  DQOs is available
from the U.S.  EPA Quality Assurance Management  Staff (QAMS) (see references  2 and
4).


2.2  PROJECT OBJECTIVES

      A statement of the project  objectives and how the objectives  are to be
attained  should  be concisely stated and sufficiently detailed to permit clear
understanding  by all  parties  involved  in the data  collection  effort.   This
includes a statement  of what problem is to be solved and  the information required


                                    ONE  - 2                           Revision  1
                                                                     July  1992

-------
in the process.   It also  includes  appropriate statements of the DQOs (i.e., the
acceptable level of uncertainty in the information).


2.3  SAMPLE COLLECTION

      Sampling  procedures,  locations,  equipment,  and sample  preservation and
handling requirements  should  be specified in  the  QAPjP.   Further  details on
quality assurance procedures for field  operations are described in Section 3 of
this chapter.  The OSW is developing policies and procedures  for sampling in a
planned  revision  of Chapter  Nine of  this  manual.   Specific procedures for
groundwater sampling are provided in Chapter Eleven of this manual.


2.4  ANALYSIS AND TESTING

      Analytes and properties of concern, analytical  and testing procedures to
be employed,  required detection limits,  and requirements for precision and bias
should be specified.  All  applicable regulatory requirements and the project DQOs
should be considered when developing the specifications.  Further details on the
procedures for analytical operations are described in  Section 4 of this chapter.


2.5  QUALITY CONTROL

      The quality  assurance program  should  address  both  field and  laboratory
activities.  Quality control procedures should be specified for estimating the
precision and bias of the data.  Recommended minimum requirements for QC samples
have been established by EPA and should be met in order to satisfy recommended
minimum criteria for acceptable data quality.   Further details  on procedures for
field and laboratory operations are described in Sections 3 and  4, respectively,
of this chapter.


2.6  PROJECT DOCUMENTATION

      Documents should be prepared and maintained in conjunction with the data
collection effort.  Project documentation should be sufficient to allow review
of all aspects of the work being performed.   The QAPjP discussed in Sections 3
and 4 is one important document that should be maintained.

      The  length  of storage  time   for project  records   should  comply  with
regulatory  requirements,  organizational  policy,   or  project  requirements,
whichever is more stringent.  It is recommended that documentation be stored for
three years from submission of the project final report.

      Documentation  should   be  secured  in   a  facility   that   adequately
addresses/minimizes  its deterioration for the  length  of time  that  it is to be
retained.  A system allowing for the expedient retrieval  of information should
exist.
                                    ONE  -  3                          Revision 1
                                                                     July 1992

-------
      Access  to  archived  information  should  be  controlled  to maintain  the
integrity  of  the  data.    Procedures  should  be  developed  to  identify  those
individuals with access to the data.


2.7  ORGANIZATION PERFORMING FIELD OR LABORATORY  OPERATIONS

      Proper design and structure  of the organization facilitates effective and
efficient transfer of information and helps to prevent important procedures from
being overlooked.

      The  organizational   structure,  functional   responsibilities,  levels  of
authority,  job  descriptions,  and  lines  of  communication   for  all  project
activities should be established and documented.  One person may  cover more than
one  organizational  function.   Each project  participant  should have  a  clear
understanding of his or her duties  and responsibilities and the relationship of
those responsibilities to the overall  data collection  effort.

      The management of each organization participating in a project involving
data collection activities should  establish that organization's operational and
QA policies.  This information should be documented in the QAPjP.  The management
should ensure that (1) the appropriate methodologies are followed as documented
in   the   QAPjPs;   (2)   personnel   clearly   understand   their  duties   and
responsibilities;  (3)  each  staff member  has access  to  appropriate  project
documents;  (4)  any  deviations  from the QAPjP  are communicated  to  the project
management  and documented; and  (5) communication  occurs between  the field,
laboratory, and project management, as specified in the QAPjP.  In addition, each
organization  should  ensure that  their activities do not  increase  the risk to
humans or the  environment  at or about the project location.  Certain projects may
require specific policies or a Health and Safety Plan to provide this assurance.

      The management of the participating field or laboratory organization should
establish personnel qualifications and training requirements for the project.
Each person participating in  the  project  should  have  the education, training,
technical  knowledge,  and  experience, or  a combination  thereof,  to  enable that
individual to perform assigned functions.   Training  should be  provided  for each
staff  member   as  necessary to perform  their  functions   properly.    Personnel
qualifications  should be  documented  in   terms of education,  experience,  and
training,   and   periodically   reviewed   to  ensure   adequacy  to   current
responsibilities.

      Each  participating  field organization  or laboratory organization should
have a  designated QA  function (i.e., a  team or individual  trained in QA) to
monitor  operations  to  ensure  that  the equipment,  personnel,   activities,
procedures, and  documentation conform with the QAPjP.   To  the extent possible,
the QA monitoring function should be entirely separate  from, and  independent of,
personnel  engaged in  the work being  monitored.    The  QA function should be
responsible for  the QA review.
                                    ONE - 4                         Revision 1
                                                                     July 1992

-------
      2.7.1  Performance Evaluation

      Performance evaluation studies are used to measure the performance of the
laboratory on  unknown  samples.   Performance evaluation samples  are typically
submitted to the laboratory as blind samples by an independent outside source.
The  results  are  compared to  predetermined acceptance  limits.    Performance
evaluation samples can  also  be  submitted to the laboratory as part of the QA
function during internal assessment of laboratory performance.  Records of all
performance evaluation  studies should be maintained by the laboratory.  Problems
identified through participation  in performance evaluation studies should be
immediately investigated and corrected.

      2.7.2  Internal Assessment by QA Function

      Personnel performing field and laboratory activities are responsible for
continually monitoring  individual compliance with  the QAPjP.   The QA function
should review procedures,  results  and calculations  to  determine compliance with
the  QAPjP.   The  results  of this  internal assessment  should be  reported to
management with requirements for a plan to correct observed deficiencies.

      2.7.3  External Assessment

      The field and laboratory  activities may be reviewed by personnel external
to the  organization.   Such an assessment  is an  extremely  valuable method for
identifying overlooked problems.   The results of the external assessment should
be submitted  to management with requirements  for a  plan  to  correct observed
deficiencies.

      2.7.4  On-Site Evaluation

      On-site evaluations may be conducted as part of both internal  and external
assessments.   The  focus of  an on-site  evaluation  is  to evaluate  the degree of
conformance of project activities with the applicable QAPjP.  On-site evaluations
may  include, but  are not  limited to,  a  complete  review of facilities, staff,
training, instrumentation, procedures,  methods, sample collection,  analyses, QA
policies and procedures  related to the generation of environmental  data.  Records
of each evaluation should include the date of the evaluation, location, the areas
reviewed,  the  person performing  the  evaluation,  findings and  problems,  and
actions recommended and  taken to resolve problems.  Any problems identified that
are  likely to affect   data  integrity   should  be  brought  immediately  to  the
attention of management.

            2.7.4.1  Field Activities

      The review of field  activities should be  conducted by one or more persons
knowledgeable  in  the activities being  reviewed  and  include  evaluating,  at a
minimum, the following  subjects:

      Completeness  of   Field  Reports   --  This  review determines  whether  all
      requirements for  field activities  in the QAPjP  have  been fulfilled, that
      complete  records  exist for each  field activity, and  that the procedures

                                    ONE - 5                         Revision 1
                                                                     July 1992

-------
      specified  in  the  QAPjP  have  been  implemented.    Emphasis  on  field
      documentation  will  help assure sample integrity and  sufficient technical
      information  to   recreate  each  field  event.     The  results  of  this
      completeness check should be documented,  and  environmental  data affected
      by incomplete  records should be identified.

      Identification of Valid Samples -- This review involves interpretation and
      evaluation of the field records  to  detect problems  affecting  the repre-
      sentativeness  of environmental  samples.   Examples  of  items  that  might
      indicate potentially invalid  samples  include improper  well  development,
      improperly screened wells,  instability of pH or conductivity, and collec-
      tion of volatiles  near  internal  combustion  engines.  The  field records
      should be evaluated against the QAPjP and  SOPs.  The reviewer should docu-
      ment the sample validity and  identify the environmental  data  associated
      with any poor  or incorrect field work.

      Correlation of  Field Test  Data  --  This  review  involves  comparing  any
      available results of field  measurements obtained by more than one method.
      For example,  surface geophysical methods should  correlate with  direct
      methods  of site  geologic  characterization  such  as   lithologic  logs
      constructed during drilling operations.

      Identification of Anomalous Field Test Data -- This review identifies any
      anomalous field  test data.   For example, a water temperature for one well
      that is  5 degrees higher  than  any  other well  temperature in  the  same
      aquifer should  be  noted.   The  reviewer should evaluate the  impact  of
      anomalous field  measurement results  on the associated environmental data.

      Validation of Field Analyses  --  This  review  validates  and  documents all
      data from  field analysis that  are  generated in  situ  or from  a mobile
      laboratory as  specified in  Section 2.7.4.2.  The reviewer should document
      whether the QC checks meet the acceptance  criteria, and whether corrective
      actions were  taken  for any analysis  performed when  acceptance criteria
      were exceeded.

            2.7.4.2   Laboratory Activities

      The review of laboratory data should be conducted  by one or more persons
knowledgeable in laboratory activities  and include evaluating,  at  a minimum, the
following subjects:

      Completeness of  Laboratory Records --  This review determines whether: (1)
      all samples and  analyses required by  the  QAPjP  have been processed, (2)
      complete records exist for  each  analysis  and the  associated QC samples,
      and that (3) the procedures specified  in the QAPjP have been implemented.
      The  results  of  the  completeness   check  should  be   documented,   and
      environmental  data affected by incomplete records  should be identified.

      Evaluation of Data with  Respect  to  Detection and  Quantitation Limits --
      This review compares analytical results to required  quantitation limits.
      Reviewers should document instances where detection or quantitation limits

                                    ONE -  6                          Revision 1
                                                                     July 1992

-------
      exceed  regulatory  limits,   action  levels,   or  target  concentrations
      specified in the QAPjP.

      Evaluation of Data with  Respect to Control Limits -- This review compares
      the   results  of QC and  calibration  check samples to control  criteria.
      Corrective  action  should be  implemented for data  not within  control
      limits.  The reviewer should check that corrective action reports, and the
      results   of reanalysis,  are   available.   The  review  should  determine
      whether  samples associated with  out-of-control QC data  are identified in
      a written  record  of the data review, and whether an assessment  of the
      utility  of such analytical results  is recorded.

      Review of Holding Time Data  -- This review compares sample holding times
      to those required by the QAPjP,  and notes all  deviations.

      Review of Performance Evaluation (PE) Results  -- PE study results can be
      helpful  in evaluating the impact of  out-of-control conditions. This review
      documents  any  recurring trends or  problems   evident in PE studies  and
      evaluates their effect on environmental  data.

      Correlation of  Laboratory Data  --  This  review determines whether  the
      results  of data obtained  from related laboratory tests,  e.g.,  Purgeable
      Organic  Halides (POX) and Volatile  Organics,  are documented,  and whether
      the significance of any differences is discussed in the  reports.

      2.7.5  QA Reports

      There should be periodic reporting  of pertinent  QA/QC information to the
project management to allow assessment of  the  overall  effectiveness  of the QA
program.  There are three major types  of  QA reports  to project management:

      Periodic Report on Key QA Activities -- Provides  summary  of key QA activi-
      ties during the  period,  stressing measures that are  being taken to improve
      data  quality;  describes  significant  quality   problems  observed  and
      corrective actions  taken; reports  information regarding any changes in
      certification/accreditation status; describes involvement in resolution of
      quality  issues  with  clients  or  agencies; reports  any QA organizational
      changes;  and  provides  notice of the distribution of revised  documents
      controlled by the QA organization  (i.e.,  procedures).

      Report on Measurement Quality  Indicators  -- Includes the assessment of QC
      data gathered over the period, the  frequency  of analyses repeated due to
      unacceptable QC performance,  and, if  possible,  the reason  for  the unac-
      ceptable performance and corrective action taken.

      Reports  on QA Assessments --  Includes the results of the assessments and
      the plan  for  correcting identified  deficiencies;  submitted  immediately
      following any internal  or external  on-site evaluation or upon  receipt of
      the results of any performance evaluation studies.
                                   ONE - 7                          Revision 1
                                                                     July 1992

-------
3.0  FIELD OPERATIONS

      The field  operations  should be  conducted  in such  a  way as  to  provide
reliable information  that  meets the DQOs.   To achieve this,  certain  minimal
policies and procedures  should be implemented.  The  OSW is considering revisions
of Chapter Nine and Eleven of this manual.  Supplemental information and guidance
is available in  the  RCRA Ground-Water Monitoring Technical Enforcement Guidance
Document (TEGD)  (Reference  3).   The project documentation  should  contain the
information specified below.


3.1  FIELD LOGISTICS

      The QAPjP should describe  the type(s)  of  field operations to be performed
and the  appropriate  area(s) in  which  to  perform  the  work.   The  QAPjP should
address ventilation,  protection  from extreme weather and temperatures, access to
stable power, and provision for water and gases of required  purity.

      Whenever practical, the  sampling  site  facilities should be examined prior
to the start of work to ensure  that all  required items are  available.  The actual
area of sampling should be examined to ensure that trucks, drilling equipment,
and personnel have adequate access to the site.

      The determination  as to whether sample  shipping is necessary should be made
during planning  for  the project.   This need is established  by evaluating the
analyses to be performed, sample holding times,  and location  of  the  site and the
laboratory.  Shipping or transporting  of samples to a laboratory should be done
within a timeframe such that recommended holding times are met.

      Samples should be  packaged, labelled, preserved (e.g., preservative added,
iced,  etc.),  and documented  in an area  which is free  of  contamination and
provides for secure storage. The level  of  custody and whether sample storage is
needed should be addressed  in the QAPjP.

      Storage areas for solvents, reagents,  standards, and reference materials
should  be   adequate  to   preserve  their identity,  concentration,   purity,  and
stability prior to use.

      Decontamination of sampling  equipment may be performed  at  the  location
where  sampling occurs,  prior  to going to the  sampling site,  or in designated
areas near the sampling  site.  Project documentation should specify  where and how
this work is accomplished.  If decontamination  is to be done  at the site, water
and  solvents of appropriate   purity  should  be   available.    The method  of
accomplishing decontamination,  including the required materials, solvents, and
water purity  should be  specified.

      During  the sampling process and  during on-site or in situ analyses, waste
materials are sometimes  generated.  The method for storage and disposal  of these
waste  materials  that  complies  with  applicable   local,   state  and  Federal
regulations should be specified. Adequate facilities should  be  provided for the
collection and storage of all wastes, and these facilities should be operated so

                                    ONE -  8                          Revision  1
                                                                     July 1992

-------
as  to  minimize  environmental  contamination.    Waste  storage  and  disposal
facilities should comply with applicable federal,  state, and local regulations.

      The location  of long-term and short-term  storage for field records, and the
measures to ensure the integrity of the data should be specified.


3.2  EQUIPMENT/INSTRUMENTATION

      The equipment, instrumentation, and supplies at the sampling site should
be specified and  should be appropriate to accomplish the activities planned.  The
equipment and  instrumentation  should  meet  the requirements of specifications,
methods, and procedures as specified in the QAPjP.


3.3  OPERATING PROCEDURES

      The QAPjP  should describe or make reference to all field activities that
may affect data quality.  For routinely performed  activities, standard operating
procedures (SOPs) are often prepared to ensure consistency  and  to save time and
effort in preparing QAPjPs.  Any deviation  from an established  procedure during
a  data  collection  activity  should be documented.   The procedures  should  be
available for  the  indicated  activities, and should  include,  at a minimum, the
information described below.

      3.3.1  Sample Management

      The numbering and labeling system, chain-of-custody procedures, and how the
samples  are to  be tracked  from  collection   to  shipment  or  receipt by the
laboratory should be specified.  Sample management procedures should also specify
the  holding times,  volumes  of sample required  by  the laboratory,  required
preservatives, and shipping requirements.

      3.3.2  Reagent/Standard Preparation

      The procedures describing how to prepare standards and reagents should be
specified.  Information concerning specific grades of materials used in reagent
and standard preparation,  appropriate glassware and containers for preparation
and storage, and labeling and record keeping for  stocks and dilutions should be
included.

      3.3.3  Decontamination

      The procedures describing decontamination  of  field  equipment before and
during  the  sample  collection process  should  be  specified.   These procedures
should include cleaning materials used, the order  of washing and rinsing with the
cleaning materials, requirements for protecting or covering cleaned equipment,
and procedures for disposing of cleaning materials.
                                    ONE -  9                         Revision 1
                                                                     July 1992

-------
      3.3.4  Sample Collection

      The  procedures  describing  how  the  sampling  operations  are  actually
performed  in  the  field should be specified.   A simple reference  to standard
methods is not sufficient,  unless  a  procedure  is performed exactly as described
in the published method.   Methods from  source documents  published  by the EPA,
American Society for Testing  and  Materials, U.S. Department  of  the Interior,
National  Water  Well  Association,  American   Petroleum  Institute,  or  other
recognized organizations with  appropriate expertise should be used, if possible.
The procedures for sample collection should include at least the following:

   •  Applicability of the procedure,

   •  Equipment required,

   •  Detailed  description of  procedures  to  be followed  in collecting  the
      samples,

   •  Common problems encountered and corrective actions to be followed, and

   •  Precautions to be taken.

      3.3.5  Field Measurements

      The  procedures  describing  all  methods used in the field  to  determine a
chemical or physical  parameter  should be  described  in  detail.  The procedures
should address criteria from  Section 4,  as  appropriate.

      3.3.6  Equipment Calibration And Maintenance

      The  procedures  describing  how  to  ensure  that  field  equipment  and
instrumentation  are  in  working order  should be  specified.    These describe
calibration  procedures and schedules,  maintenance procedures  and schedules,
maintenance  logs,  and  service  arrangements  for  equipment.    Calibration  and
maintenance of field equipment and instrumentation  should be in accordance with
manufacturers'  specifications or  applicable test  specifications and should be
documented.

      3.3.7  Corrective Action

      The  procedures describing how to identify and correct deficiencies  in the
sample collection  process  should  be specified.  These should include specific
steps  to  take  in  correcting   deficiencies   such  as  performing  additional
decontamination  of  equipment,  resampling,  or additional  training  of field
personnel.  The procedures should specify that each corrective action  should be
documented with a description  of the deficiency and the  corrective action  taken*
and  should include the person(s) responsible  for  implementing the corrective
action.
                                   ONE  -  10                         Revision 1
                                                                     July  1992

-------
      3.3.8  Data Reduction and Validation

      The procedures describing how to compute results from field measurements
and to review and validate these data  should be specified.  They should include
all formulas  used to calculate results  and  procedures  used  to independently
verify that field measurement results are correct.

      3.3.9  Reporting

      The procedures describing the process for reporting the results of field
activities should be specified.

      3.3.10 Records Management

      The  procedures  describing  the  means for  generating,  controlling,  and
archiving  project-specific records  and  field  operations  records  should  be
specified.  These procedures should detail record generation and  control and the
requirements for record retention, including type, time, security, and retrieval
and disposal authorities.

      Project-specific records  relate to field work performed  for a project.
      These records may  include correspondence, chain-of-custody records, field
      notes, all reports issued as a result of the work,  and procedures used.

      Field operations records  document overall field operations  and may include
      equipment performance and maintenance logs, personnel  files, general field
      procedures, and corrective action reports.

      3.3.11 Waste Disposal

      The  procedures  describing the  methods  for disposal  of waste materials
resulting from field operations should be specified.


3.4  FIELD QA AND QC REQUIREMENTS

      The  QAPjP  should  describe how  the  following   elements of the field QC
program will be implemented.

      3.4.1  Control Samples

      Control  samples  are QC  samples that are  introduced into a  process to
monitor the performance of the system.  Control samples,  which may include blanks
(e.g.,  trip,  equipment,  and  laboratory),   duplicates,  spikes,   analytical
standards, and reference materials,  can  be used in different phases of the data
collection process beginning with sampling and continuing through transportation,
storage, and analysis.

      Each  day of sampling, at least one field duplicate and one equipment
rinsate should be collected for each matrix sampled.  If this frequency is not
appropriate for the sampling equipment and method, then the appropriate changes

                                   ONE -  11                         Revision 1
                                                                     July 1992

-------
should be^TeaHy identified in  the  QAPjP.    When  samples are collected  for
volatile organic analysis, a trip blank  is also  recommended  for each  day that
samples are collected.   In addition, for  each sampling batch (20 samples of one
matrix type), enough  volume should be collected for at least one sample so as to
allow the laboratory  to prepare one matrix spike and either  one matrix duplicate
or one matrix spike duplicate for each analytical method employed.   This means
that the following control samples are recommended:

      •Field duplicate (one per day per matrix type)
      •Equipment rinsate  (one per day per matrix type)
      •Trip blank (one per day,  volatile organics only)
      •Matrix spike (one per batch [20 samples of each matrix type])
      •Matrix duplicate or matrix spike duplicate (one per batch)

Additional control samples may be necessary in order to assure data quality to
meet the project-specific DQOs.

      3.4.2  Acceptance Criteria

      Procedures  should  be in  place for establishing acceptance  criteria  for
field activities described in the QAPjP.  Acceptance criteria may be qualitative
or  quantitative.    Field  events  or  data that  fall  outside of  established
acceptance criteria may indicate a problem with the sampling process that should
be investigated.

      3.4.3  Deviations

      All deviations from plan  should be documented as to the  extent of,  and
reason  for,  the  deviation.   Any  activity not  performed  in  accordance with
procedures or QAPjPs  is considered a deviation  from plan.   Deviations from plan
may or may not affect data quality.

      3.4.4  Corrective Action

      Errors, deficiencies, deviations, certain field events,  or data that fall
outside  established  acceptance  criteria should  be investigated.   In  some in-
stances,  corrective  action may be  needed  to  resolve the  problem  and restore
proper  functioning to  the system.  The investigation of  the  problem and any
subsequent corrective action taken should be documented.

      3.4.5  Data Handling

      All  field  measurement  data  should  be  reduced  according  to protocols
described or referenced in the QAPjP.  Computer programs  used  for data reduction
should be validated before use and verified on a regular basis.  All  information
used in the calculations  should be recorded to enable reconstruction of the final
result at a  later date.

      Data should be  reported in accordance with the requirements of the end-user
as described  in the  QAPjP.


                                   ONE  -  12                         Revision  1
                                                                     July 1992

-------
3.5  QUALITY ASSURANCE REVIEW

      The QA Review consists of internal and external  assessments to ensure that
QA/QC procedures  are  in  use and  to  ensure that field staff  conform  to these
procedures.  QA review should  be  conducted  as deemed appropriate and necessary.


3.6  FIELD RECORDS

      Records provide  the direct evidence and support for the necessary technical
interpretations, judgments, and discussions concerning project activities. These
records, particularly those that  are anticipated to be used as  evidentiary data,
should directly support current or ongoing  technical studies and activities and
should provide  the  historical  evidence needed  for later  reviews and analyses.
Records should  be legible, identifiable, and retrievable and protected against
damage, deterioration, or loss.  The discussion in this section (3.6) outlines
recommended procedures for record  keeping.   Organizations  which conduct field
sampling  should develop appropriate  record  keeping procedures which satisfy
relevant technical and legal requirements.

      Field records generally consist of bound  field notebooks with prenumbered
pages,  sample  collection forms,  personnel  qualification and  training forms,
sample  location maps,  equipment maintenance and  calibration  forms,  chain-of-
custody forms,  sample  analysis  request forms,  and field  change request forms.
All records should  be written in indelible ink.

      Procedures for reviewing, approving,  and  revising field  records should be
clearly defined, with the lines of authority included.  It is recommended that
all documentation errors  should be corrected by  drawing a single line through the
error  so  it  remains  legible  and  should  be  initialed  by  the  responsible
individual, along  with  the  date of change.  The  correction  should  be written
adjacent to the error.

      Records should include (but are not limited to) the following:

      Calibration Records & Traceability of Standards/Reagents  -- Calibration is
      a reproducible  reference  point to which all  sample  measurements can be
      correlated.   A  sound  calibration program should include  provisions for
      documentation of frequency, conditions, standards, and records reflecting
      the  calibration  history  of a  measurement  system.   The  accuracy of the
      calibration standards is important because all data will be in reference
      to  the  standards  used.    A  program for  verifying  and  documenting the
      accuracy  of all  working standards against primary grade standards should
      be routinely followed.

      Sample Collection -- To  ensure  maximum  utility of the sampling effort and
      resulting data, documentation  of the  sampling  protocol, as  performed in
      the field, is essential.  It is  recommended that sample collection records
      contain,  at  a minimum,  the names of  persons conducting the  activity,
      sample  number,  sample  location, equipment  used,  climatic  conditions,
      documentation of adherence to  protocol,  and unusual observations.   The

                                   ONE - 13                         Revision 1
                                                                     July 1992

-------
      actual  sample  collection  record  is usually one of the following:  a bound
      field notebook with prenumbered pages, a pre-printed  form,  or digitized
      information on a computer tape or  disc.

      Chain-of-Custody Records  -- The chain-of-custody involving the possession
      of samples from  the  time they are obtained  until  they are  disposed  or
      shipped off-site  should be documented as  specified in the QAPjP and should
      include the following information:   (1)  the  project name;  (2) signatures
      of samplers;  (3)  the  sample number, date and time of collection, and grab
      or composite sample designation; (4)  signatures of individuals involved in
      sample  transfer; and  (5)  if  applicable,  the air bill or  other shipping
      number.

      Maps and Drawings --  Project planning documents and reports often contain
      maps.  The maps  are  used to  document the location of sample collection
      points  and monitoring wells  and  as  a means of  presenting environmental
      data.  Information used to prepare maps and drawings is normally obtained
      through field  surveys,  property surveys, surveys  of monitoring  wells,
      aerial  photography or photogrammetric mapping.   The final, approved maps
      and/or  drawings should have a  revision number and date and should be sub-
      ject to the same controls as  other project  records.

      QC Samples --  Documentation for generation of QC  samples, such as trip and
      equipment rinsate blanks, duplicate  samples, and any field spikes should
      be maintained.

      Deviations --  All deviations  from  procedural   documents  and  the  QAPjP
      should  be recorded in the site logbook.

      Reports -- A  copy of  any  report issued  and  any  supporting documentation
      should  be retained.
4.0  LABORATORY OPERATIONS

      The laboratory should conduct its operations in such a way as to provide
reliable information.  To achieve  this, certain minimal policies and procedures
should be implemented.


4.1  FACILITIES

      The QAPjP  should  address  all  facility-related  issues  that  may  impact
project  data  quality.     Each  laboratory  should  be  of  suitable  size  and
construction to facilitate the proper conduct of the analyses.  Adequate bench
space or working  area per analyst  should be provided.   The  space requirement per
analyst depends on the equipment or apparatus that is  being utilized, the number
of samples that the analyst is expected to  handle  at any one time,  and the number
of operations that are to be  performed  concurrently by a single analyst.  Other
issues to be considered include,  but  are not  limited to, ventilation, lighting,


                                   ONE -  14                         Revision 1
                                                                     July 1992

-------
control of dust and drafts,  protection  from extreme temperatures, and access to
a source of stable power.

      Laboratories should be designed  so  that  there  is  adequate separation of
functions to ensure  that no laboratory activity has an  adverse  effect  on the
analyses. The laboratory may require specialized facilities such as a perchloric
acid hood or glovebox.

      Separate space for  laboratory operations and  appropriate ancillary support
should be provided,  as needed,  for  the performance  of  routine  and specialized
procedures.

      As  necessary  to  ensure  secure   storage  and  prevent contamination  or
misidentification, there should be adequate facilities for receipt and storage
of samples.   The level   of  custody  required  and any special requirements for
storage such as refrigeration should be described in planning documents.

      Storage areas for reagents, solvents, standards,  and reference materials
should  be adequate  to   preserve  their identity,  concentration, purity,  and
stability.

      Adequate facilities should be provided for the collection and storage of
all wastes,  and these facilities should be operated so as to minimize environ-
mental contamination. Waste storage and disposal facilities should comply with
applicable federal, state, and local regulations.

      The location of long-term and short-term storage of laboratory records and
the measures to ensure the integrity of the data should be specified.


4.2  EQUIPMENT/INSTRUMENTATION

      Equipment and instrumentation  should meet the requirements and specifica-
tions of  the  specific test methods and other procedures as specified  in the
QAPjP.  The  laboratory should maintain  an  equipment/instrument description list
that includes the manufacturer, model number,  year of  purchase, accessories, and
any modifications, updates, or upgrades that have been made.


4.3  OPERATING PROCEDURES

      The QAPjP should describe or make reference to all laboratory activities
that may affect data quality. For routinely performed activities,  SOPs are often
prepared to ensure consistency  and  to save time and effort in preparing QAPjPs.
Any deviation from  an established procedure  during  a data collection activity
should be documented.  It is recommended  that procedures  be available for the
indicated activities, and  include, at a minimum,  the  information described
below.
                                   ONE - 15                         Revision 1
                                                                     July 1992

-------
      4.3.1  Sample Management

      The procedures describing the receipt,  handling,  scheduling,  and storage
of samples should be specified.

      Sample Receipt and Handling -- These procedures describe the  precautions
      to be used in opening sample  shipment  containers  and  how to  verify that
      chain-of-custody has been maintained,  examine  samples  for damage,  check
      for  proper preservatives  and  temperature,  and  log  samples   into  the
      laboratory sample streams.

      Sample Scheduling --  These procedures  describe the sample scheduling in
      the laboratory and  includes procedures used  to ensure  that holding time
      requirements are met.

      Sample Storage -- These procedures describe the  storage  conditions for all
      samples, verification and documentation of daily storage temperature, and
      how to  ensure that  custody  of the samples  is maintained while  in  the
      laboratory.

      4.3.2  Reagent/Standard Preparation

      The procedures describing how to prepare standards and  reagents  should be
specified.  Information concerning  specific grades  of materials  used in reagent
and standard preparation,  appropriate glassware and containers for preparation
and storage, and labeling  and recordkeeping for stocks and dilutions  should be
included.

      4.3.3  General Laboratory Techniques

      The procedures describing all  essentials of laboratory operations that are
not addressed elsewhere should be specified.   These techniques  should include,
but are not limited to, glassware cleaning procedures, operation of analytical
balances, pipetting techniques, and use of volumetric glassware.

      4.3.4  Test Methods

      Procedures  for  test methods  describing  how  the  analyses are  actually
performed in the laboratory should be  specified.  A simple  reference to standard
methods is not sufficient, unless the  analysis is performed exactly as  described
in  the  published method.  Whenever methods  from  SW-846  are not appropriate,
recognized methods  from source documents published by the EPA,  American Public
Health Association (APHA), American Society for Testing and Materials  (ASTM), the
National  Institute  for  Occupational Safety  and  Health  (NIOSH),   or  other
recognized organizations with  appropriate expertise should be used, if  possible.
The documentation  of the  actual laboratory  procedures  for analytical methods
should  include  the  following:

      Sample  Preparation  and  Analysis Procedures  --  These include applicable
      holding time, extraction, digestion, or preparation steps  as appropriate
      to  the  method; procedures for determining the appropriate  dilution to

                                    ONE - 16                         Revision  1
                                                                      July  1992

-------
      analyze;  and  any  other  information  required  to  perform the  analysis
      accurately and consistently.

      Instrument Standardization -- This includes concentration(s) and frequency
      of analysis of  calibration  standards, linear  range  of the method,  and
      calibration acceptance criteria.

      Sample Data -- This includes recording requirements and documentation in-
      cluding sample identification  number,  analyst, data verification, date of
      analysis and verification,  and computational  method(s).

      Precision and  Bias -- This  includes all  analytes for which the method is
      applicable and the conditions  for use of this information.

      Detection  and Reporting Limits  --  This  includes all  analytes  in  the
      method.

      Test-Specific  QC  --  This  describes QC  activities  applicable to  the
      specific test  and references  any applicable  QC  procedures.

      4.3.5  Equipment Calibration  and Maintenance

      The procedures  describing  how to  ensure  that  laboratory  equipment  and
instrumentation  are in working order  should  be  specified.   These  procedures
include  calibration  procedures  and   schedules,  maintenance  procedures  and
schedules,  maintenance logs, service arrangements  for all equipment, and spare
parts available in-house.  Calibration and maintenance of laboratory equipment
and instrumentation  should be in  accordance with manufacturers'  specifications
or applicable test specifications  and should be documented.

      4.3.6  QC

      The type,  purpose, and  frequency  of QC samples  to  be analyzed  in  the
laboratory and the acceptance criteria  should be specified.  Information should
include  the  applicability  of  the   QC  sample  to  the  analytical process,  the
statistical  treatment  of  the data, and the responsibility of laboratory  staff and
management in generating and using the  data.  Further details on development of
project-specific QC  protocols are  described in Section 4.4.

      4.3.7  Corrective Action

      The procedures describing how  to  identify and correct deficiencies in the
analytical  process should be specified.  These should include specific steps to
take in  correcting  the deficiencies such as preparation of  new standards  and
reagents,  recalibration   and  restandardization of equipment,  reanalysis  of
samples,  or  additional   training   of  laboratory  personnel  in  methods  and
procedures.   The procedures  should specify that each corrective action  should be
documented with a description of the deficiency and  the corrective action taken,
and should  include  the person(s)  responsible for  implementing  the  corrective
action.
                                   ONE - 17                         Revision 1
                                                                     July 1992

-------
      4.3.8  Data Reduction and Validation

      The procedures describing how to review and  validate  the  data should be
specified.  They should  include procedures  for  computing  and interpreting the
results from QC samples, and independent procedures to verify  that the analytical
results are reported correctly.  In addition, routine procedures used to monitor
precision and bias, including evaluations of reagent, equipment rinsate, and trip
blanks,  calibration  standards, control  samples,   duplicate  and matrix  spike
samples, and surrogate  recovery,  should  be detailed in the  procedures.   More
detailed validation procedures  should be performed  when  required in the contract
or QAPjP.

      4.3.9  Reporting

      The procedures describing the process for reporting the  analytical results
should be specified.

      4.3.10 Records Management

      The  procedures  describing  the  means  for generating,  controlling,  and
archiving laboratory records should  be specified.   The procedures should detail
record generation and control, and  the requirements for record retention, includ-
ing type, time, security, and retrieval  and disposal authorities.

      Project-specific  records may  include correspondence,  chain-of-custody
      records,  request for analysis,  calibration data records, raw and finished
      analytical and QC data, data reports, and procedures used.

      Laboratory operations records may include laboratory notebooks, instrument
      performance logs and maintenance logs  in bound notebooks with prenumbered
      pages; laboratory  benchsheets;  software  documentation;  control  charts;
      reference material certification; personnel  files; laboratory procedures;
      and corrective action reports.

      4.3.11 Waste Disposal

      The procedures describing the  methods  for  disposal of chemicals including
standard and reagent solutions, process waste, and samples should be specified.


4.4  LABORATORY QA AND QC PROCEDURES

      The  QAPjP should  describe  how the  following required elements  of the
laboratory QC program are to be implemented.

      4.4.1  Method Proficiency

      Procedures should be  in  place  for demonstrating proficiency with each
analytical  method   routinely  used in  the  laboratory.   These  should include
procedures for demonstrating the  precision  and  bias  of the method as performed
by  the  laboratory   and  procedures for  determining  the method detection limit

                                   ONE -  18                         Revision  1
                                                                     July 1992

-------
(MDL).  All terminology, procedures and frequency of determinations associated
with the laboratory's establishment of the MDL and the reporting  limit should be
well-defined  and  well-documented.    Documented  precision,  bias,  and  MDL
information should be maintained for all methods performed in the laboratory.

      4.4.2  Control Limits

      Procedures should be in place for establishing and updating control limits
for  analysis.   Control  limits should  be  established  to  evaluate laboratory
precision and bias based  on the analysis of control  samples.  Typically, control
limits for bias are  based  on the  historical  mean recovery plus  or minus three
standard deviation units, and control limits for precision range from zero (no
difference between duplicate control  samples)  to the historical mean relative
percent difference plus three standard deviation units.  Procedures  should be in
place for monitoring  historical  performance and should include graphical (control
charts) and/or tabular presentations of the data.

      4.4.3  Laboratory  Control Procedures

      Procedures should be in place for demonstrating that the laboratory is in
control during each  data collection activity.   Analytical  data  generated with
laboratory control samples that fall within prescribed  limits are judged to be
generated while the  laboratory was  in control.  Data generated with laboratory
control samples that fall outside the established control limits are judged to
be generated during  an  "out-of-control"  situation.   These data  are considered
suspect and should be repeated or reported with qualifiers.

      Laboratory  Control Samples  --   Laboratory  control  samples  should  be
      analyzed for each  analytical  method  when  appropriate  for  the method.   A
      laboratory control sample consists of either  a control matrix  spiked with
      analytes representative  of  the target  analytes or a certified reference
      material.

      Laboratory control sample(s) should be analyzed with each batch of samples
      processed to verify that  the precision  and bias of the analytical process
      are  within  control  limits.    The   results  of  the  laboratory  control
      sample(s) are  compared to control  limits  established for  both precision
      and bias to determine usability of the data.

      Method Blank -- When appropriate for the method,  a method  blank should be
      analyzed with  each batch of samples processed  to assess contamination
      levels in the laboratory. Guidelines should  be in place for accepting or
      rejecting data based on the level of contamination in the blank.

      Procedures should be in place for documenting  the effect of the matrix on
method performance.  When appropriate for the method,  there should  be at least
one matrix spike and either one matrix duplicate or  one  matrix spike duplicate
per analytical  batch. Additional control samples may be necessary to assure data
quality to meet the  project-specific DQOs.
                                   ONE - 19                         Revision 1
                                                                     July 1992

-------
      Matrix-Specific Bias  --  Procedures should be in place for determining the
      bias of the method due  to  the matrix.   These procedures  should  include
      preparation and analysis of matrix spikes,  selection  and use of surrogates
      for organic methods,  and the method of standard  additions  for metal  and
      inorganic methods.  When the concentration of the  analyte in the sample is
      greater than 0.1%,  no spike is necessary.

      Matrix-Specific Precision -- Procedures should  be  in  place for determining
      the precision  of the method  for a specific  matrix.   These  procedures
      should  include  analysis  of  matrix   duplicates  and/or  matrix  spike
      duplicates.  The frequency  of use of these techniques  should be based on
      the DQO for the data  collection activity.

      Matrix-Specific  Detection  Limit -- Procedures  should be  in  place  for
      determining the MDL for a specific matrix type  (e.g., wastewater treatment
      sludge, contaminated  soil,  etc).

      4.4.4  Deviations

      Any activity not performed in accordance with laboratory  procedures or
QAPjPs is considered  a deviation  from plan.  All deviations from plan should be
documented as to the extent of, and reason for,  the  deviation.

      4.4.5  Corrective Action

      Errors, deficiencies, deviations, or laboratory events or data that fall
outside  of  established acceptance criteria  should  be  investigated.   In some
instances, corrective  action may be  needed to  resolve  the problem and restore
proper functioning to  the analytical system.   The investigation of the problem
and any subsequent corrective action taken should be documented.

      4.4.6  Data Handling

      Data resulting  from the  analyses  of samples should be reduced according to
protocols described  in the  laboratory  procedures.   Computer programs used for
data reduction should  be validated before use and verified on a regular basis.
All information  used in  the calculations (e.g.,  raw data,  calibration files,
tuning records,  results of standard additions,  interference check results, and
blank- or background-correction protocols) should be recorded in  order to enable
reconstruction  of the final   result   at  a  later date.    Information on  the
preparation  of the sample  (e.g., weight  or volume of sample used, percent dry
weight  for  solids,  extract  volume,  dilution   factor  used)   should  also be
maintained in order to enable reconstruction of  the final result at a  later date.

      All data should be  reviewed by a  second  analyst or supervisor according to
laboratory  procedures to ensure that  calculations  are correct  and to detect
transcription errors.  Spot checks  should  be  performed  on  computer calculations
to  verify program validity.   Errors detected  in  the  review process should be
referred  to  the  analyst(s)  for corrective action.  Data should be reported in
accordance with  the  requirements of the end-user.  It is recommended that the
supporting documentation include at a  minimum:

                                   ONE -  20                          Revision 1
                                                                     July  1992

-------
      Laboratory name and address.

      Sample  information  (including  unique  sample  identification,  sample
      collection date and time, date of  sample  receipt,  and  date(s)  of sample
      preparation and analysis).

      Analytical results  reported with  an  appropriate number  of  significant
      figures.

      Detection limits that  reflect dilutions, interferences, or correction for
      equivalent dry weight.

      Method reference.

      Appropriate QC results (correlation with sample batch should be traceable
      and documented).

      Data qualifiers with appropriate references and narrative on the quality
      of the results.
4.5  QUALITY ASSURANCE REVIEW

      The QA review consists of internal and external  assessments to ensure that
QA/QC procedures are in use and to ensure that laboratory  staff conform to these
procedures.  QA review should  be  conducted  as deemed appropriate and necessary.


4.6  LABORATORY RECORDS

      Records provide the direct evidence and support for the necessary technical
interpretations,  judgements,  and  discussions  concerning project  activities.
These records, particularly those that are anticipated to  be used as evidentiary
data, should directly support  technical  studies and activities, and provide the
historical evidence  needed for later  reviews  and  analyses.   Records should be
legible,   identifiable,   and   retrievable,  and  protected   against   damage,
deterioration,  or  loss.    The  discussion  in  this   section  (4.6)  outlines
recommended procedures for record  keeping.   Organizations  which  conduct field
sampling  should  develop  appropriate  record  keeping procedures which  satisfy
relevant technical and legal  requirements.

      Laboratory records generally consist of bound notebooks with prenumbered
pages,  personnel  qualification  and training forms,  equipment  maintenance and
calibration forms,  chain-of-custody forms,  sample  analysis  request forms, and
analytical change request forms.  All records should be written in indelible ink.

      Procedures for reviewing, approving, and revising laboratory records should
be clearly  defined,  with  the  lines of  authority  included.   Any  documentation
errors should be corrected by drawing a single line through the error so that it
remains legible and should be  initialed by the responsible  individual, along with
the date of change.  The correction is written adjacent to the error.

                                   ONE - 21                         Revision 1
                                                                     July 1992

-------
      Strip-chart recorder printouts should be signed by the person who performed
the instrumental  analysis.   If corrections need  to be made  in computerized data,
a system parallel to the corrections for handwritten  data should be in place.

      Records of sample management should be available  to permit the re-creation
of an analytical  event for review in the case of  an audit or investigation of a
dubious result.

      Laboratory records should include, at least, the following:

      Operating Procedures -- Procedures should be available to those performing
      the task outlined.    Any  revisions  to  laboratory  procedures  should  be
      written, dated,  and distributed to  all  affected individuals  to ensure
      implementation  of  changes.   Areas  covered by   operating  procedures are
      given in Sections 3.3 and 4.3.

      Quality Assurance Plans -- The QAPjP should be  on file.

      Equipment Maintenance  Documentation  -- A history of the maintenance record
      of each  system serves as  an  indication  of the  adequacy  of maintenance
      schedules and parts  inventory.  As appropriate, the maintenance guidelines
      of the  equipment  manufacturer should be followed.   When  maintenance  is
      necessary,  it  should be  documented  in  either  standard  forms or  in
      logbooks.  Maintenance procedures  should be clearly defined and written
      for each measurement  system and required support equipment.

      Proficiency -- Proficiency information on all compounds reported  should be
      maintained and  should include (1) precision; (2) bias;  (3) method detec-
      tion limits; (4) spike recovery, where applicable; (5) surrogate recovery,
      where  applicable;  (6) checks on  reagent purity, where  applicable; and
      (7) checks on glassware cleanliness, where  applicable.

      Calibration Records &  Traceability of Standards/Reagents  -- Calibration is
      a  reproducible  reference  point  to which all sample measurements can  be
      correlated.   A sound  calibration program  should  include  provisions for
      documenting frequency, conditions, standards, and records reflecting the
      calibration  history  of   a  measurement  system.   The  accuracy  of the
      calibration standards is  important because  all   data will be in  reference
      to  the  standards  used.    A program  for verifying  and  documenting the
      accuracy  and  traceability of all working  standards against appropriate
      primary grade standards or the highest quality standards available should
      be routinely  followed.

      Sample Management -- All  required records pertaining to sample management
      should  be maintained  and updated  regularly.   These  include  chain-of-
      custody  forms,  sample receipt forms, and sample disposition records.

      Original  Data  -- The raw data  and  calculated results  for all  samples
      should be maintained  in laboratory notebooks, logs,  benchsheets,  files or
      other sample tracking or data entry forms.   Instrumental output  should be
      stored  in  a computer  file  or a hardcopy  report.

                                   ONE  -  22                         Revision  1
                                                                      July 1992

-------
      QC Data  -- The  raw data and  calculated results  for  all  QC  and field
      samples and standards should be maintained in the manner described in the
      preceding  paragraph.   Documentation should allow  correlation  of sample
      results with associated  QC  data.   Documentation  should also  include the
      source and lot numbers of standards for traceability.  QC samples include,
      but are not limited  to, control samples, method  blanks,  matrix spikes, and
      matrix spike duplicates.

      Correspondence -- Project correspondence can provide evidence supporting
      technical interpretations.   Correspondence pertinent to  the project should
      be kept and placed  in the project files.

      Deviations -- All deviations from procedural  and planning documents should
      be recorded  in  laboratory  notebooks.   Deviations  from QAPjPs  should be
      reviewed  and approved  by   the  authorized  personnel  who performed  the
      original technical   review or by their designees.

      Final Report --  A copy of any  report issued and any supporting documenta-
      tion should be retained.
5.0  DEFINITIONS
      The following terms are defined for use in this document:
ACCURACY
BATCH:
BIAS:
The closeness of agreement between an observed value  and
an accepted  reference  value.   When  applied  to  a set  of
observed  values,  accuracy will  be  a combination  of  a
random  component  and of  a common  systematic  error  (or
bias) component.

A group of samples which behave  similarly  with respect  to
the sampling or the testing procedures being  employed  and
which are processed as a unit (see Section  3.4.1  for  field
samples and Section 4.4.3 for laboratory samples).  For QC
purposes, if the number of samples in a group is greater
than 20, then each group of 20 samples or less will all  be
handled as a separate batch.

The deviation due to matrix effects of the measured  value
(xs -  xu) from a known spiked amount.  Bias  can be  assessed
by comparing  a measured value  to  an accepted reference
value in a sample  of known concentration  or by determining
the recovery of a known  amount of contaminant spiked into
a sample (matrix spike). Thus, the bias  (B) due  to matrix
effects based on a matrix spike is calculated as:
                        where:
                                    B = (x. -  xu  )  - K
                                   ONE - 23
                                               Revision  1
                                               July  1992

-------
BLANK:

CONTROL SAMPLE:
DATA QUALITY
OBJECTIVES (DQOs)
DATA VALIDATION:
DUPLICATE:


EQUIPMENT BLANK:

EQUIPMENT RINSATE:
ESTIMATED
QUANTITATION
LIMIT (EQL):
                                    xs  =  measured  value  for  spiked  sample,
                                    xu  = measured value for unspiked sample, and
                                    K  =  known value of the spike in the sample.

                      Using the following equation yields the percent recovery
              %R =  100  (xs - xj/  K

see Equipment Rinsate, Method Blank, Trip Blank.

A  QC  sample  introduced  into a  process to  monitor the
performance of the system.

A  statement  of  the overall  level of uncertainty that a
decision-maker  is  willing to accept  in results derived
from environmental  data (see reference  2, EPA/QAMS, July
16, 1986).   This  is qualitatively distinct from quality
measurements such as precision, bias, and detection limit.
The process of evaluating the available data against the
project DQOs  to  make sure that  the  objectives are met.
Data  validation  may  be  very
depending on  project  DQOs.   The
will include analytical  results,
data, and may also include field
                                                       rigorous,   or  cursory,
                                                       available  data reviewed
                                                      field QC data and lab QC
                                                       records.
see  Matrix  Duplicate,
Duplicate.

see Equipment Rinsate.
Field  Duplicate,   Matrix  Spike
A  sample  of analyte-free  media  which has  been used  to
rinse  the sampling  equipment.    It  is  collected  after
completion of decontamination and prior to sampling.  This
blank is useful  in documenting adequate decontamination of
sampling equipment.

The  lowest  concentration that can  be reliably  achieved
within specified limits  of precision  and  accuracy during
routine  laboratory  operating  conditions.   The  EQL  is
generally  5  to  10  times the MDL.    However,  it may  be
nominally chosen within these  guidelines  to  simplify data
reporting.      For   many   analytes  the   EQL   analyte
concentration is selected as the  lowest non-zero standard
in the calibration curve.  Sample EQLs are highly matrix-
dependent.  The  EQLs in  SW-846 are  provided  for guidance
and may not always be achievable.
                                   ONE - 24
                                               Revision  1
                                                July  1992

-------
FIELD DUPLICATES:
LABORATORY CONTROL
SAMPLE:
MATRIX:


MATRIX DUPLICATE:


MATRIX SPIKE:
MATRIX SPIKE
DUPLICATES:
METHOD BLANK:
METHOD DETECTION
LIMIT (MDL):
Independent  samples  which  are  collected  as  close  as
possible to the same point  in  space  and time.  They are
two separate samples taken from the same source,  stored in
separate containers, and  analyzed independently.   These
duplicates are useful  in documenting  the precision of the^
sampling process.

A known matrix spiked with compound(s)  representative of
the target analytes.   This is used to document laboratory
performance.

The component or substrate (e.g.,  surface water, drinking
water) which contains the  analyte of interest.

An intralaboratory split sample which is used  to document
the precision of a method  in a given sample matrix.

An aliquot of sample spiked with a known concentration of
target  analyte(s).   The spiking  occurs prior to sample
preparation  and  analysis.   A  matrix  spike  is  used to
document the bias of a method in a given sample matrix.

Intralaboratory  split  samples   spiked  with   identical
concentrations of target analyte(s).  The  spiking occurs
prior to sample preparation and analysis.  They are used
to document the precision  and bias of a method in a given
sample matrix.

An analyte-free matrix to  which all reagents are added in
the  same  volumes  or  proportions  as  used   in  sample
processing.   The  method blank  should be carried through
the complete sample preparation and analytical procedure.
The  method  blank  is  used to  document  contamination
resulting from the analytical process.

For  a method  blank  to be  acceptable   for  use  with the
accompanying  samples,  the  concentration in the blank of
any  analyte of concern should not  be  higher  than the
highest of either:

(l)The method detection limit,  or

(2)Five percent of the regulatory limit for that analyte,
or

(3)Five  percent  of  the  measured concentration  in  the
sample.

The  minimum concentration  of   a  substance that  can  be
measured and reported with 99% confidence that  the analyte
concentration is greater than zero and  is determined from
                                   ONE - 25
                                              Revision  1
                                               July  1992

-------
analysis of a sample  in  a  given  matrix type  containing
the analyte.

For  operational  purposes,  when  it  is  necessary  to
determine  the   MDL  in  the matrix,  the  MDL should  be
determined by multiplying the appropriate one-sided 99% t-
statistic  by  the  standard  deviation  obtained  from  a
minimum of three analyses of a matrix spike containing the
analyte of interest at a concentration three  to five times
the estimated MDL, where the t-statistic  is obtained from
standard references  or the table below.
No. of samples;            t-statistic
      3                       6.96
      4                       4.54
      5                       3.75
      6                       3.36
      7                       3.14
      8                       3.00
      9                       2.90
     10                       2.82

Estimate the MDL as follows:
Obtain the concentration value that corresponds to:

a) an  instrument  signal/noise  ratio within the range of
2.5 to 5.0, or

b) the region  of  the standard  curve where  there is  a
significant change  in  sensitivity (i.e.,  a break  in  the
slope of the standard curve).

Determine the variance (S2) for each analyte as follows:
                       n-1
where Xi  =  the  ith measurement of the variable x
and x = the average value of x;
              ONE -  26                         Revision  1
                                                July 1992

-------
                      Determine the standard deviation (s) for each analyte as
                      follows:
                                            1/2
                      Determine the MDL for each  analyte as follows:
ORGANIC-FREE
REAGENT WATER:
PRECISION:
                                    MDL
                                           '(n-1, o = .99)
                                (S)
where t(n_.,     99. is the one-sided t-statistic appropriate
for the number'or samples used to determine (s), at the 99
percent level.

For  volatiles,  all  references  to water  in  the  methods
refer to water in  which an  interferant is  not observed at
the method detection limit of the compounds  of interest.
Organic-free reagent water  can be generated by passing tap
water through a  carbon  filter bed containing about 1 pound
of activated carbon.   A water purification system may be
used   to   generate   organic-free    deionized    water.
Organic-free reagent water  may also be prepared by boiling
water for 15 minutes and, subsequently, while maintaining
the temperature  at 90°C, bubbling a contaminant-free inert
gas through the water  for 1 hour.

For  semivolatiles  and  nonvolatiles,   all  references  to
water  in  the  methods  refer  to  water  in  which   an
interferant is not observed at  the method  detection limit
of the compounds of interest.   Organic-free reagent water
can  be  generated  by passing tap  water through  a carbon
filter bed containing  about 1 pound of activated carbon.
A  water purification  system  may be used  to  generate
organic-free deionized water.

The  agreement   among   a  set  of  replicate   measurements
without  assumption  of  knowledge  of  the   true  value.
Precision  is  estimated by  means  of  duplicate/replicate
analyses.  These samples should contain concentrations of
analyte above the MDL, and may  involve the use of matrix
spikes.  The most commonly used estimates of precision are
the relative standard  deviation  (RSD)  or  the coefficient
of variation (CV),

               RSD = CV =  100  S/x,
                                   ONE - 27
                                               Revision  1
                                                July  1992

-------
PROJECT:
QUALITY ASSURANCE
PROJECT PLAN
(QAPJP):
RCRA:

REAGENT BLANK:

REAGENT GRADE:




REAGENT WATER:
REFERENCE MATERIAL:
SPLIT SAMPLES:
STANDARD ADDITION:
STANDARD CURVE:
                     wjiere:
                      x  =  the  arithmetic mean of the Xj measurements,  and  S =
                     variance;  and  the relative percent difference (RPD) when
                     only  two  samples  are  available.
               RPD =  100
                                                   - x2)/{(Xl + x2)/2}].
Single or  multiple data collection  activities that are
related through the same planning sequence.

An orderly assemblage of detailed procedures designed to
produce  data of  sufficient  quality to  meet  the data
quality  objectives   for   a  specific  data   collection
activity.

The Resource Conservation and Recovery Act.

See Method Blank.

Analytical  reagent (AR) grade,  ACS  reagent  grade, and
reagent  grade  are  synonymous  terms  for  reagents  which
conform to the current specifications of the  Committee on
Analytical Reagents of the American Chemical  Society.

Water that  has  been generated by any method which  would
achieve  the  performance  specifications  for ASTM Type II
water.    For organic  analyses,   see the  definition of
organic-free reagent water.

A material containing known quantities of  target analytes
in solution  or  in a homogeneous  matrix.   It  is used to
document the bias of the analytical  process.

Aliquots  of sample  taken   from  the  same  container and
analyzed  independently.    In  cases  where  aliquots of
samples are impossible to obtain, field  duplicate samples
should be taken for the matrix duplicate analysis.   These
are usually taken after mixing or compositing and are used
to document  intra-  or  interlaboratory precision.

The practice of adding  a known amount of  an  analyte to  a
sample  immediately prior to  analysis.   It is typically
used to  evaluate  interferences.

A plot of concentrations of  known analyte standards versus
the  instrument  response  to  the analyte.    Calibration
standards are prepared by successively diluting a standard
solution  to produce  working standards which  cover the
working  range  of  the instrument.   Standards  should be
prepared  at the  frequency  specified in the  appropriate
                                   ONE - 28
                                               Revision 1
                                                July 1992

-------
                      section.   The  calibration  standards should  be prepared
                      using  the same type of  acid  or solvent and  at the same
                      concentration  as  will  result  in  the  samples  following
                      sample preparation.  This  is applicable to  organic and
                      inorganic chemical  analyses.

SURROGATE:            An   organic  compound  which  is  similar  to  the  target
                      analyte(s)  in chemical  composition  and behavior  in the
                      analytical  process, but which  is not  normally found in
                      environmental  samples.

TRIP BLANK:           A sample  of analyte-free media  taken from the laboratory
                      to   the  sampling  site   and  returned to  the  laboratory
                      unopened.   A trip  blank  is used to document contamination
                      attributable  to shipping and field  handling  procedures.
                      This type of blank is useful in documenting contamination
                      of  volatile organics samples.


6.0  REFERENCES

1.   Interim  Guidelines  and  Specifications   for  Preparing  Quality  Assurance
     Project Plans, QAMS-005/80, December 29,  1980,  Office  of  Monitoring Systems
     and Quality Assurance,  ORD,  U.S.  EPA, Washington,  DC  20460.

2.   Development of Data Quality Objectives, Description of Stages I  and II, July
     16, 1986, Quality Assurance Management Staff, ORD,  U.S. EPA, Washington, DC
     20460.

3.   RCRA  Ground-Water Monitoring  Technical  Enforcement Guidance  Document,
     September, 1986, Office of Waste Programs Enforcement.   OSWER,  U.S. EPA,
     Washington, DC,  20460.

4.   DQO  Training  Software, Version  6.5,  December,  1988,  Quality  Assurance
     Management Staff, ORD,  U.S.  EPA,  Washington, DC  20460.

5.   Preparing  Perfect Project  Plans,   EPA/600/9-89/087,  October  1989,  Risk
     Reduction Engineering Laboratory (Guy Simes),  Cincinnati OH.

6.   ASTM Method D 1129-77,  Specification for Reagent Water.   1991  Annual Book
     of ASTM Standards.  Volume 11.01 Water  and  Environmental Technology.

7.   Generation of Environmental  Data  Related to  Waste  Management Activities
     (Draft).  February 1989.   ASTM.
                                   ONE - 29                         Revision 1
                                                                     July 1992

-------
                                     INDEX

Accuracy  1, 13, 22, 23*, 24
Batch  12, 19, 21, 23*
Bias  2, 3, 17-20, 22, 23*-25, 28
Blank  11, 12, 14, 18-20, 23*, 24, 25, 28, 29
   Equipment Rinsate  11, 12,  14,  18,  24*
   Method Blank  19, 24, 25*, 28
   Reagent Blank  28*
   Trip Blank  12, 18, 24,  29*
Chain-of-Custody  9, 11, 13,  14,  18,  21,  22
Control Chart  18, 19
Control Sample  11, 12,  18,  19, 23, 24*
Data Quality Objectives  (DQO)  1-3, 8, 12,  19,  20,  24", 28
Decision-maker  2, 24
Duplicate  11, 12, 14, 18-20,  23,  24", 25, 27,  28
   Field Duplicate  11,  12,  24, 25", 28
   Matrix Duplicate  12, 19,  20,  24,  25", 28
   Matrix Spike Duplicate   12, 19,  20, 23,  24,  25"
Equipment Blank  11, 24"
Equipment Rinsate  11, 12,  14, 18,  24"
Estimated Quantitation Limit  (EQL)  24"
Field Duplicate  12, 24, 25", 28
Laboratory Control Sample   19, 25"
Matrix  11, 12, 18-20, 23-25*, 26-28
Matrix Duplicate  12, 19, 20,  24,  25", 28
Matrix Spike  12, 18-20, 23,  25",  26, 27
Matrix Spike Duplicate   12,  19, 20, 23,  24,  25"
Method Blank  19, 24, 25",  28
Method Detection Limit (MDL)  18-20, 22,  24,  25*-27
Organic-Free Reagent Water   27",  28
Precision  1-3, 17-20, 22,  24, 25,  27*,  28
Project  1-5, 7, 8, 11-14,  17-19,  21,  23, 24,  28*
Quality Assurance Project Plan (QAPjP)  1-9,  11, 12,  14, 15,  18, 20, 22, 23, 28*
RCRA  1, 8, 28"
Reagent Blank  28*
Reagent Grade  28*
Reagent Water  27, 28*
Reference Material  8, 11,  15, 18,  19, 28*
Split Samples  25, 28"
Standard Addition  20, 28*
Standard Curve  26, 28"
Surrogate  18, 20, 22, 29"
Trip Blank  12, 18, 24,  29*
   Definition of term.
                                    ONE -  30                         Revision 1
                                                                      July 1992

-------
                                 CHAPTER TWO

                        CHOOSING THE CORRECT PROCEDURE
2.1  PURPOSE
      This chapter  aids  the analyst  in  choosing the appropriate  methods for
samples, based upon sample matrix and the analytes to be determined.

      2.1.1  Trace Analysis vs. Macroanalvsis

      The methods presented in SW-846 were  designed  through  sample sizing and
concentration procedures  to address the problem of "trace" analyses  (<1000 ppm),
and have been developed for an  optimized working  range.  These methods are also
applicable to "minor"  (1000 ppm -  10,000 ppm) and "major"  (>10,000 ppm) analyses,
as well as to  "trace"  analyses,  through  use of appropriate sample preparation
techniques that result in analyte  concentration within that optimized range. Such
sample preparation techniques include:

      1) adjustment of size of sample prepared for analysis,
      2) adjustment of injection volumes,
      3) dilution or concentration of sample,
      4) elimination of concentration steps prescribed for "trace" analyses,
      5) direct injection (of samples to be analyzed for volatile constituents).

      The performance  data presented in each  of these methods were generated from
"trace" analyses, and may  not  be  applicable to  "minor"  and "major" analyses.
Generally, extraction efficiency improves as concentration increases.

CAUTION:    Care  should  be taken  when analyzing  samples  for trace analyses
            subsequent  to  analysis  of   concentrated   samples   due  to  the
            possibility of contamination.

      2.1.2  Choice  of Apparatus  and  Preparation  of Reagents

      Since many  types and sizes of  glassware  and supplies  are  commercially
available, and since  it  is  possible  to prepare  reagents  and  standards in many
different ways, those  specified in these methods  may be replaced by any similar
types as long  as  this  substitution does  not affect the  overall  quality of the
analyses.


2.2  REQUIRED  INFORMATION

      In order to  choose the  correct  combination  of  methods  to  form  the
appropriate analytical procedure, some basic information is required.
                                   TWO  -  1                      Revision 2
                                                            September 1994

-------
      2.2.1   Physical  State(s)  of Sample

      The phase  characteristics of the sample must be known.  There are several
general categories of phases in which the sample may be categorized:

              Aqueous                 Oil and Organic  Liquid
              Sludges                 Solids
              Multiphase Samples      EP and TCLP Extracts
              Ground Water

      2.2.2  Analytes

      Analytes are divided into classes based on  the determinative methods which
are used to identify and quantify them.  Table 2-1 lists the organic analytes of
SVI-846 methods,  Table 2-2 lists the analytes that may be prepared using Method
3650,  and Table  2-3  lists  the analytes that  are  collected  from  stack  gas
effluents  using VOST  methodology.   Tables  2-4 through  2-31 list the target
analytes of each organic determinative method.  Some of the analytes  appear on
more than one table,  as  they may be determined using  any  of several methods.
Table 2-32 indicates which methods are applicable to  inorganic target analytes.

      2.2.3  Detection Limits  Required

      Regulations may require  a specific sensitivity or detection limit for an
analysis, as in  the determination of analytes  for  the Toxicity Characteristic
(TC) or  for  delisting  petitions.  Drinking water detection  limits,  for those
specific organic and metallic  analytes covered by the National Interim Primary
Drinking Water Standards, are desired in the analysis of ground water.

      2.2.4  Analytical Objective

      Knowledge  of  the analytical objective will  be useful  in  the  choice of
aliquoting procedures and in the selection of a determinative method.  This is
especially true when  the sample  has more than one  phase.   Knowledge of  the
analytical objective may not be possible  or desirable at all management levels,
but  that  information  should   be  transmitted  to  the analytical  laboratory
management  to  ensure  that  the  correct  techniques   are  being applied  to  the
analytical effort.

      2.2.5  Detection and Monitoring

      The strategy for detection of compounds  in environmental  or process samples
may be contrasted with the strategy for monitoring samples.  Detection samples
define initial  conditions.   When there is little information available about the
composition of the sample source, e.g.,  a well or process stream, mass  spectral
identification of organic analytes leads to fewer false positive results.  Thus,
the most practical form of  detection  for  organic analytes, given the analytical
requirements, is mass spectral  identification.   The  choice  of  technique  for
metals   is   governed   by   the   detection  limit  requirements  and   potential
interferents.

      Monitoring samples, on the other hand,  are analyzed to confirm existing and
on-going  conditions,  tracking   the presence  or  absence of constituents  in an


                                    TWO  - 2                      Revision 2
                                                            September  1994

-------
environmental  or  process  matrix.    In  well  defined  and  stable  analytical
conditions and matrices less compound-specific detection modes may be used.

      2.2.6  Sample Containers. Preservations, and Holding Times

      Appropriate sample containers,  sample preservation techniques, and sample
holding times for aqueous matrices are listed  in Table 2-33, near the  end of this
chapter.   Similar  information may  be  found  in  Table 3-1  of  Chapter  Three
(inorganic analytes) and Table 4-1 of Chapter Four (organic analytes).  Samples
must be extracted/analyzed within  the specified holding times for the results to
be considered  reflective  of total concentrations.  Analytical  data generated
outside of the specified holding times must be considered to be minimum values
only.  Such data may be used to demonstrate that a waste is hazardous where it
shows the concentration  of a constituent to be above the regulatory threshold but
cannot be used to demonstrate that a waste is not hazardous.
2.3  IMPLEMENTING THE GUIDANCE

      The  choice  of the  appropriate  sequence  of  methods  depends  on  the
information required and on the experience of the analyst.  Figure 2-1 summarizes
the organic analysis options available.  Appropriate selection is confirmed by
the quality control  results.  The use of the recommended procedures, whether they
are approved or mandatory, does not release the analyst from demonstrating the
correct execution of the method.

      2.3.1  Extraction and Sample Preparation Procedures

      Methods for preparing organic analytes  are  shown in Table  2-34.   Method
3500 and associated methods should be consulted for further details on preparing
the sample for analysis.

            2.3.1.1  Aqueous Samples

            The choice of a preparative method depends on the sample.  Methods
      3510  and  3520 may  be  used for  extraction  of the  semivolatile  organic
      compounds.  Method 3510,  a  separatory  funnel  extraction,  is appropriate
      for samples which will  not form a persistent emulsion interphase between
      the sample and the extraction solvent.   The  formation of an emulsion that
      cannot  be  broken  up   by mechanical   techniques  will  prevent  proper
      extraction  of the  sample.   Method  3520,   a liquid-liquid  continuous
      extraction, may be used for any aqueous  sample; this method will minimize
      emulsion formation.

                  2.3.1.1.1  Basic or Neutral Extraction of Semivolatiles

                  The solvent extract obtained by  performing either Method 3510
            or 3520 at  a neutral or  basic pH  will  contain the  compounds  of
            interest.  Refer to Table 1 in the extraction methods (3510 and/or
            3520) for guidance  on the  pH  requirements  for extraction prior to
            analysis.
                                   TWO  - 3                      Revision 2
                                                            September 1994

-------
            2.3.1.1.2  Acidic Extraction of Phenols and Acids

            The extract obtained by performing either Method 3510 or 3520
      at a pH less than or equal  to 2  will  contain the phenols and acid
      extractables.

      2.3.1.2  Solid Samples

      Soxhlet  (Methods 3540  and  3541)  and  ultrasonic  (Method  3550)
extractions are used with solid samples.  Consolidated samples should be
ground  finely  enough   to  pass  through  a  1   mm  sieve.    In  limited
applications, waste  dilution  (Method  3580)  may  be  used  if  the  entire
sample is soluble in the specified solvent.

      Methods 3540% and  3541  and  3550 are neutral-pH extraction techniques
and therefore, depending on the analysis requirements, acid-base partition
cleanup (Method 3650) may be necessary.  Method 3650 will  only be needed
if chromatographic interferences are severe enough to prevent detection of
the analytes of interest.  This  separation  will  be most important if a GC
method is chosen  for analysis of the sample.   If GC/MS is used,  the ion
selectivity  of   the  technique  may   compensate  for  chromatographic
interferences.

      2.3.1.3  Oils and Organic Liquids

      Method 3580, waste dilution,  may be  used  and the resultant sample
analyzed directly by GC or  GC/MS.   To avoid overloading  the analytical
detection system, care must be exercised to ensure that proper dilutions
are made.  Method 3580 gives guidance  on performing waste dilutions.

      To remove  interferences,  Method 3611 may be performed  on  an oil
sample directly,  without prior sample  preparation.

      Method 3650 is the  only other preparative  procedure  for oils and
other  organic  liquids.   This  procedure  is a  back extraction  into an
aqueous  phase.    It  is  generally introduced as  a cleanup procedure for
extracts rather than as a preparative  procedure.   Oils generally have a
high concentration of semivolatile compounds and, therefore, preparation
by Method 3650 should be done on a relatively small aliquot of the sample.
Generally,  extraction  of 1  ml  of  oil will be sufficient to  obtain  a
saturated aqueous phase and avoid emulsions.

      2.3.1.4  Sludge Samples

      There  is no set ratio of liquid  to solid which enables the analyst
to  determine which  of  the  three  extraction methods  cited  is  the most
appropriate.   If the  sludge  is  an organic sludge  (solid  material and
organic  liquid,  as opposed  to an  aqueous  sludge),  the sample should be
handled as a multiphase sample.

      Determining the   appropriate  methods for  analysis  of  sludges is
complicated  because  of the lack of precise definition of  sludges with
respect to the relative percent of liquid and solid components.  They may
be  classified into three categories  but with appreciable  overlap.

                              TWO - 4                     Revision  2
                                                      September 1994

-------
            2.3.1.4.1  Liquids

            Use of Method 3510 or Method 3520 may be applicable to sludges
      that  behave like  and have  the consistency  of aqueous  liquids.
      Ultrasonic  extraction  (Method  3550)  and  Soxhlet  (Method  3540)
      procedures  will,  most likely,  be   ineffective  because  of  the
      overwhelming presence of the liquid aqueous phase.

            2.3.1.4.2  Solids

            Soxhlet  (Methods  3540 and 3541)  and ultrasonic  extraction
      (Method 3550) will  be more effective when  applied to sludge samples
      that resemble solids.  Samples may be dried or centrifuged to form
      solid  materials  for  subsequent  determination  of  semivolatile
      compounds.

            Using Method  3650, Acid-Base Partition Cleanup, on the extract
      may be necessary, depending on whether  chromatographic interferences
      prevent determination of the analytes of interest.

            2.3.1.4.3  Emulsions

            Attempts should be made to break up and separate the phases of
      an emulsion. Several techniques are effective in breaking emulsions
      or separating the phases of emulsions.

      1.  Freezing/thawing:  Certain emulsions will separate if exposed to
          temperatures below 0°C.

      2.  Salting out:  Addition of a salt to make the aqueous phase of an
          emulsion too  polar to  support  a less polar  phase  promotes
          separation.

      3.  Centrifugation:     Centrifugal   force  may  separate  emulsion
          components  by density.

      4.  Addition of water  or  ethanol:     Emulsion  polymers  may  be
          destabilized when a preponderance of the aqueous phase is added.

            If techniques for breaking emulsions  fail, use Method 3520.
      If the emulsion  can be broken, the different phases (aqueous, solid,
      or organic liquid)  may then be analyzed separately.

      2.3.1.5  Multiphase Samples

      Choice of  the procedure for  aliquoting  multiphase samples  is very
dependent on the objective of the analysis.   With a  sample in which some
of the phases tend to  separate rapidly,  the percent weight or volume of
each phase  should be calculated  and  each  phase  should be  individually
analyzed for the required analytes.

      An alternate approach is to  obtain  a homogeneous sample and attempt
a single analysis on the combination of phases.  This  approach will give


                              TWO  - 5                      Revision 2
                                                      September 1994

-------
      no Information on the abundance of the analytes In the Individual  phases
      other than what can be implied by solubility.

            A third alternative is to select phases  of interest and to analyze
      only those  selected phases.   This tactic must  be consistent  with  the
      sampling/analysis objectives or it will yield insufficient information for
      the time and resources expended.   The phases  selected should be compared
      with Figure 2-1 and Tables 2-34 through 2-36  for further guidance.

      2.3.2  Cleanup Procedures

      Each  category  in  Table  2-35,  Cleanup  of  Organic  Analyte  Extracts,
corresponds to one of the possible  determinative methods available  in the manual.
Cleanups employed are determined by the analytes of interest within the extract.
However, the necessity of performing cleanup may also depend upon the matrix from
which the extract was developed.   Cleanup of a sample  may  be  done  exactly as
instructed in  the  cleanup method  for  some of the  analytes.   There  are some
instances when  cleanup using one  of the methods may  only proceed  after  the
procedure is  modified to optimize  recovery and separation.   Several  cleanup
techniques may be possible for  each analyte  category.  The information provided
is not  meant  to imply that any or  all  of these methods must  be  used for  the
analysis  to  be  acceptable.    Extracts  with  components which interfere with
spectral  or  chromatographic determinations  are expected  to  be  subjected  to
cleanup procedures.

      The  analyst's  discretion  must   determine  the   necessity  for  cleanup
procedures, as there are no clear cut criteria for indicating their use.  Method
3600 and associated methods should be consulted for further details on extract
cleanup.

      2.3.3  Determinative Procedures

      The determinative methods for organic analytes have been divided  into three
categories, shown in Table 2-36:   gas chromatography/mass spectrometry (GC/MS);
specific detection methods,  i.e.,  gas chromatography  (GC); and high performance
liquid  chromatography (HPLC).   This division  is  intended to  help  an analyst
choose which determinative method  will apply.  Under  each analyte column,  SW-846
method numbers have been indicated,  if appropriate,  for the determination  of the
analyte.  A  blank  has been  left if no  chromatographic determinative method is
available.

      Generally, the MS procedures are more specific but less sensitive than the
appropriate gas chromatographic/specific detection method.

      Method  8000  gives  a  general   description   of   the  technique  of  gas
chromatography.  This method should be  consulted  prior  to application  of  any of
the gas chromatographic methods.

      Methods 8080 and 8081, for  organochlorine pesticides and polychlorinated
biphenyls, Methods 8140 and 8141, for organophosphorus  pesticides, and Me'thods
8150 and 8151,  for chlorinated herbicides,  are preferred over GC/MS because of
the  combination of  selectivity  and  sensitivity  of  the  flame  photometric,
nitrogen-phosphorus,  and  electron capture detectors.


                                    TWO - 6                     Revision 2
                                                            September 1994

-------
      Methods 8240 and 8260 are both GC/MS methods for volatile analytes.  Method
8240 uses a packed column whereas Method 8260 employs a capillary column.  Better
chromatographic separation of the  volatile compounds  may be obtained by using
Method 8260 rather than 8240.  Performance criteria will be based on Method 8260.
Method 5030 has been combined with  both Method 8240  and 8260, with which it was
used  exclusively.    A GC with  a  selective  detector is  also useful  for  the
determination of volatile organic compounds in a monitoring scenario, described
in Sec. 2.2.5.

      Methods 8250  and 8270  are both  GC/MS methods  for semi volatile analytes.
Method 8250 uses a packed column whereas Method 8270  employs  a capillary column.
Better chromatographic separation of the semi volatile  compounds may be obtained
by using Method 8270 rather  than 8250.  Performance criteria will  be based on
Method 8270.
2.4  CHARACTERISTICS

      Figure 2-2 outlines a sequence for determining  if  a waste exhibits one or
more of the characteristics of a hazardous waste.

      2.4.1  EP and TCLP extracts

      The leachate obtained from using either the EP (Figure 2-3A) or the TCLP
(Figure 2-3B)  is  an aqueous  sample,  and therefore, requires  further solvent
extraction prior to the analysis of semivolatile compounds.

      The TCLP leachate is solvent extracted with methylene chloride at  a pH > 11
and at a pH <2 by either Method 3510 or 3520.   Method  3510 should be used unless
the formation  of emulsions between the  sample and the  solvent  prevent proper
extraction.  If this problem is encountered,  Method 3520 should be employed.

      The solvent extract obtained by performing either Method 3510 or 3520 at
a basic or neutral pH will  contain the base/neutral compounds of interest.  Refer
to the  specific determinative  method for guidance on the  pH requirements for
extraction prior to analysis.

      Due  to  the  high  concentration of  acetate  in the  TCLP extract,  it  is
recommended that purge-and-trap be used to introduce the volatile sample into the
gas chromatograph.


2.5  GROUND WATER

      Appropriate analysis schemes for the determination of analytes in ground
water are  presented in Figures 2-4A, 2-4B, and 2-4C.  Quantitation limits for the
metallic  analytes  should correspond  to the drinking  water limits which are
available.
                                   TWO  -  7                      Revision 2
                                                            September 1994

-------
      2.5.1 Special Techniques for Metal Analvtes

      All  atomic  absorption  analyses  should  employ appropriate  background
correction systems whenever spectral  interferences  could  be present.   Several
background  correction techniques  are  employed in  modern  atomic  absorption
spectrometers.  Matrix modification can complement background correction in some
cases.  Since no approach to interference correction  is completely effective in
all cases, the analyst should attempt to verify the adequacy of correction.   If
the interferant is known  (e.g.  high concentrations of iron  in the determination
of selenium), accurate analyses of synthetic  solutions of the interferant (with
and without analyte)  could establish the efficacy of  the background correction.
If the nature of the interferant is not established, good agreement of analytical
results using  two substantially different wavelengths could substantiate  the
adequacy of the background correction.

      To  reduce matrix  interferences,  all  graphite  furnace atomic absorption
(GFAA) analyses should be performed using techniques which maximize an isothermal
environment within the furnace cell.  Data indicate  that  two such  techniques,
L'vov platform and the Delayed Atomization Cuvette (DAC), are equivalent in this
respect, and produce  high quality results.

      All furnace atomic  absorption analysis should be carried out using the best
matrix modifier for the analysis.  Some  examples of modifiers are listed below.
(See also the appropriate methods.)

            Element(s)           Modifier(s)

            As  and Se         Nickel  nitrate, palladium
            Pb                Phosphoric acid,  ammonium phosphate,  palladium
            Cd                Ammonium phosphate, palladium
            Sb                Ammonium nitrate, palladium
            Tl                Platinum, palladium

      The ICP calibration standards must match the acid composition and strength
of the  acids  contained in the  samples.  Acid strengths  in the ICP calibration
standards should  be stated in the raw data.

      2.5.2  Special  Techniques for Indicated Analvtes and Anions

      If  an  Auto-Analyzer is used  to  read  the  cyanide distillates,  the
spectrophotometer  must be used with  a 50 mm path length  cell.  If  a sample is
found  to  contain cyanide, the sample must  be  redistilled  a second  time  and
analyzed  to  confirm the  presence  of the cyanide.  The second distillation must
fall within the 14-day holding time.

2.6  REFERENCES

1.   Barcelona, M.J.  "TOC  Determinations in  Ground  Water"; Ground  Water 1984,
     22(1). 18-24.
                                    TWO - 8                     Revision 2
                                                            September 1994

-------
2.   Riggin,  R.;  et al.  Development  and  Evaluation of Methods for Total Organic
     Halide  and  Purqeable Organic  Halide  In  Wastewater;  U.S.  Environmental
     Protection   Agency.  Office  of  Research  and  Development.  Environmental
     Monitoring  and Support Laboratory. ORD  Publication  Offices  of Center for
     Environmental  Research Information: Cincinnati, OH, 1984; EPA-600/4-84-008.

3.   McKee,  G.;  et  al.  Determination  of Inorganic  Anions in  Water  bv Ion
     Chromatographv; (Technical  addition  to Methods  for  Chemical  Analysis of
     Water and  Wastewater,  EPA  600/4-79-020),  U.S.   Environmental  Protection
     Agency.  Environmental Monitoring and  Support  Laboratory.  ORD Publication
     Offices  of Center for Environmental Research Information:  Cincinnati, OH,
     1984; EPA-600/4-84-017.
                                   TWO  - 9                      Revision 2
                                                            September 1994

-------
                                   TABLE  2-1
            DETERMINATIVE ANALYTICAL METHODS FOR ORGANIC COMPOUNDS
Compound
Applicable Method(s)
i
Acenaphthene
Acenaphthylene
Acetaldehyde
Acetone
Acetonitrile
Acetophenone
2-Acetylami nof1uorene
1-Acetyl-2-thiourea
Acifluorfen
Acrolein (Propenal)

Acrylamide
Acrylonitrile
Alachlor
Aldicarb (Temik)
Aldicarb Sulfone
Aldrin
Allyl alcohol
Allyl chloride
2-Ami noanthraqui none
Aminoazobenzene
4-Aminobiphenyl
2-Amino-4,6-dinitrotoluene (2-Am-DNT)
4-Amino-2,6-dinitrotoluene (4-Am-DNT)
3-Ami no-9-ethylcarbazole
Anilazine
Aniline
o-Anisidine
Anthracene
Aramite
Aroclor-1016 (PCB-1016)
Aroclor-1221 (PCB-1221)
Aroclor-1232 (PCB-1232)
Aroclor-1242 (PCB-1242)
Aroclor-1248 (PCB-1248)
Aroclor-1254 (PCB-1254)
Aroclor-1260 (PCB-1260)
Aspon
Asulam
Atrazine
Azinphos-ethyl
Azinphos-methyl
Barban
Bentazon
8100, 8250/8270, 8310, 8410
8100, 8250/8270, 8310, 8410
8315
8240/8260, 8315
8240/8260
8250/8270
8270
8270
8151
8030/8031, 8240/8260, 8315,
8316
8032, 8316
8030/8031, 8240/8260, 8316
8081
8318
8318
8080/8081, 8250/8270, 8275
8240/8260
8010, 8240/8260
8270
8270
8250/8270
8330
8330
8270
8270
8250/8270
8270
8100, 8250/8270, 8310, 8410
8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8141
8321
8141
8141
8140/8141, 8270
8270
8151
                                   TWO - 10
                    Revision 2
               September  1994

-------
                                   TABLE  2-1.
                                   (Continued)
Compound
Applicable  Method(s)
Benzal chloride
Benzaldehyde
Benz(a)anthracene
Benzene
Benzidine
Benzo(b)fluoranthene
Benzo(j)f1uoranthene
Benzo(k)f1uoranthene
Benzoic acid
Benzo(g,h,i)perylene
Benzo(a)pyrene

p-Benzoquinone
Benzotrichloride
Benzyl alcohol
Benzyl benzoate
Benzyl chloride
BHC (Hexachlorocyclohexane)
a-BHC (alpha-Hexachlorocyclohexane)
/3-BHC (beta-Hexachl orocycl ohexane)
8-BHC (delta-Hexachlorocyclohexane)
•y-BHC (Lindane, gamma-Hexachlorocycl ohexane)
Bi s(2-Chloroethoxy)methane
Bis(2-Chloroethyl)ether
Bi s(2-Chloroethyl)sulfide
Bis(2-Chloroisopropyl) ether
Bis(2-Ethylhexyl) phthalate
Bolstar (Sulprofos)
Bromoacetone
Bromobenzene
Bromochloromethane
Bromodi chloromethane
4-Bromofluorobenzene
Bromoform
Bromomethane
4-Bromophenyl phenyl ether
Bromoxynil
Butanal
n-Butanol
2-Butanone (Methyl  ethyl ketone, MEK)
n-Butylbenzene
sec-Butyl benzene
tert-Butylbenzene
Butyl  benzyl phthalate
8121
8315
8100, 8250/8270,  8310,  8410
8020, 8021,  8240/8260
8250/8270
8100, 8250/8270,  8310
8100
8100, 8250/8270,  8275,  8310
8250/8270, 8410
8100, 8250/8270,  8310
8100, 8250/8270,  8275,  8310,
8410
8270
8121
8250/8270
8061
8010, 8121,  8240/8260
8120
8080/8081, 8121,  8250/8270
8080/8081, 8121,  8250/8270
8080/8081, 8121,  8250/8270
8080/8081, 8121,  8250/8270
8010, 8110,  8250/8270,  8410
8110, 8250/8270,  8410
8240/8260
8010, 8110,  8250/8270,  8410
8060/8061, 8250/8270, 8410
8140/8141
8010, 8240/8260
8010, 8021,  8260
8021, 8240/8260
8010, 8021,  8240/8260
8240/8260
8010, 8021,  8240/8260
8010, 8021,  8240/8260
8110, 8250/8270, 8410
8270
8315
8260
8015, 8240/8260
8021, 8260
8021, 8260
8021, 8260
8060/8061, 8250/8270, 8410
                                   TWO - 11
                   Revision 2
               September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
2-sec-Butyl-4,6-dinitrophenol (DNBP, Dinoseb)
Captafol
Captan
Carbaryl (Sevin)
Carbazole
Carbofuran (Furaden)
Carbon disulfide
Carbon tetrachloride
Carbophenothion (Carbofenthion)
Chloral hydrate
Chloramben
Chlordane (technical)
a-Chlordane
7-Chlordane
Chlorfenvinphos
Chloroacetonitrile
4-Chloroaniline
Chlorobenzene
Chiorobenzilate
2-Chloro-l,3-butadiene
1-Chlorobutane
Chiorodibromomethane (Dibromochloromethane)
Chloroethane
2-Chloroethanol
2-Chloroethyl vinyl ether
Chloroform
1-Chlorohexane
Chioromethane
5-Chloro-2-methylaniline
Chloromethyl methyl ether
4-Chloro-3-methylphenol
Chloroneb
3-(Chloromethyl)pyridine hydrochloride
1-Chioronaphthalene
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenol
4-Chloro-l,2-phenylenediamine
4-Chloro-l,3-phenylenediamine
4-Chlorophenyl phenyl ether
Chloroprene
3-Chloropropene
3-Chloropropionitrile
Chloropropylate
8040, 8150/8151, 8270, 8321
8081, 8270
8081, 8270
8270, 8318
8275
8270, 8318
8240/8260
8010, 8021, 8240/8260
8141, 8270
8240/8260
8151
8080, 8250/8270
8081
8081
8141, 8270
8260
8250/8270, 8410
8010, 8020, 8021, 8240/8260
8081, 8270
8260
8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8010, 8240/8260
8010, 8240/8260
8010, 8021, 8240/8260
8010, 8260
8010, 8021, 8240/8260
8270
8010
8040, 8250/8270, 8275, 8410
8081
8270
8250/8270, 8275
8120/8121, 8250/8270, 8410
8040, 8250/8270, 8275, 8410
8410
8270
8270
8110, 8250/8270, 8410
8010, 8240/8260
8260
8240/8260
8081
                                   TWO - 12
                    Revision  2
                September  1994

-------
                                   TABLE  2-1.
                                   (Continued)
Compound
Applicable  Method(s)
Chlorothalonil
2-Chlorotoluene
4-Chlorotoluene
Chlorpyrifos
Chlorpyrifos methyl
Chrysene
Coumaphos
Coumarin Dyes
p-Cresidine
o-Cresol (2-Methylphenol)
m-Cresol (3-Methylphenol)
p-Cresol (4-Methylphenol)
Cresols (Methylphenols,  Cresylic  acids)
Crotonaldehyde
Crotoxyphos
Cyclohexanone
2-Cyclohexyl-4,6-dinitrophenol
2,4-D
Dalapon
2,4-DB
DBCP
2,4-D, butoxyethanol ester
DC PA
DCPA diacid
4,4'-DDD
4,4'-DDE
4,4'-DDT
Decanal
Demeton-0,  and -S
2,4-D,ethylhexyl ester
Dial!ate
2,4-Diaminotoluene
Diazinon
Dibenz(a,h)acridine
Dibenz(a,j)acridine
Dibenz(a,h)anthracene
7H-Dibenzo(c,g)carbazole
Dibenzofuran
Dibenzo(a,e)pyrene
Dibenzo(a,h)pyrene
Dibenzo(a,i)pyrene
Dibenzothiophene
Di bromochloromethane (Chiorodi bromomethane)
1,2-Di bromo-3-chloropropane
 8081
 8021,  8260
 8010,  8021, 8260
 8140/8141
 8141
 8100,  8250/8270, 8310,  8410
 8140/8141, 8270
 8321
 8270
 8250/8270, 8410
 8270
 8250/8270, 8275, 8410
 8040
 8260,  8315
 8141,  8270
 8315
 8040,  8270
 8150/8151, 8321
 8150/8151, 8321
 8150/8151, 8321
 8081
 8321
 8081
 8151
 8080/8081, 8250/8270
 8080/8081, 8270
 8080/8081, 8250/8270
 8315
 8140/8141, 8270
 8321
 8081,  8270
 8270
 8140/8141
 8100
 8100,  8250/8270
 8100,  8250/8270, 8310
 8100
 8250/8270, 8410
 8100, 8270
 8100
 8100
 8275
8010, 8021,  8240/8260
8010, 8011,  8240/8260,  8270
                                   TWO - 13
                   Revision 2
               September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
1,2-Dibromoethane (Ethylene dibromide)
Di bromof1uoromethane
Dibromomethane
Di-n-butyl phthalate
Dicamba
Dichlone
1,2-Dichlorobenzene

1,3-Dichlorobenzene

1,4-Dichlorobenzene

3,3'-Dichlorobenzidine
3,5-Dichlorobenzoic acid
l,4-Dichloro-2-butene
cis-l,4-Dichloro-2-butene
trans-l,4-Dichloro-2-butene
Di chlorodi f1uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene (Vinylidene chloride)
ci s-1,2-Di chloroethene
trans-1,2-Dichloroethene
Dichlorofenthion
Dichloromethane (Methylene chloride)
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorprop
1,2-Dichloropropane
1,3-Dichloropropane
2,2-Di chloropropane
l,3-Dichloro-2-propanol
1,1-Dichloropropene
cis-l,3-Dichloropropene
trans-1,3-Dichloropropene
Dichlorvos  (Dichlorovos)
Dichrotophos
Dicofol
Dieldrin
1,2,3,4-Di epoxybutane
Diethyl ether
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
8010, 8011, 8021, 8240/8260
8260
8010, 8021, 8240/8260
8060/8061, 8250/8270, 8410
8150/8151, 8321
8081, 8270
8010, 8020, 8021, 8120/8121,
8250/8270, 8260, 8410
8010, 8020, 8021, 8120/8121,
8250/8270, 8260, 8410
8010, 8020, 8021, 8120/8121,
8250/8270, 8260, 8410
8250/8270
8151
8010, 8240
8260
8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8021, 8260
8010, 8021, 8240/8260
8141
8010, 8021, 8240/8260
8040, 8250/8270, 8275, 8410
8040, 8250/8270
8150/8151, 8321
8010, 8021, 8240/8260
8021, 8260
8021, 8260
8010, 8240/8260
8021, 8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8140/8141, 8270, 8321
8141, 8270
8081
8080/8081, 8250/8270
8240/8260
8015, 8260
8060/8061, 8250/8270, 8410
8270
8270
                                   TWO - 14
                   Revision 2
               September  1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
1,4-Difluorobenzene
Dihydrosaffrole
Dimethoate
3,3'-Dimethoxybenzidine
Dimethylaminoazobenzene
2,5-Dimethylbenzaldehyde
7,12-Dimethylbenz(a)anthracene
3,3'-Dimethylbenzidine
a,a-Dimethylphenethylamine
2,4-Dimethylphenol
Dimethyl phthalate
Dinitrobenzene
1,2-Dinitrobenzene
1,3-Dinitrobenzene (1,3-DNB)
1,4-Dinitrobenzene
4,6-Dinitro-2-methylphenol
2,4-Dinitrophenol
2,4-Dinitrotoluene (2,4-DNT)

2,6-Dinitrotoluene (2,6-DNT)
Dinocap
Dinoseb (2-sec-Butyl-4,6-dinitrophenol, DNBP)
Di-n-octyl phthalate
Di-n-propyl phthalate
Dioxacarb
1,4-Dioxane
Dioxathion
Diphenylamine
5,5-Di phenylhydantoi n
1,2-Diphenylhydrazine
Disperse Blue 3
Disperse Blue 14
Disperse Brown 1
Disperse Orange 3
Disperse Orange 30
Disperse Red 1
Disperse Red 5
Disperse Red 13
Disperse Red 60
Disperse Yellow 5
Disulfoton
Endosulfan I
Endosulfan II
Endosulfan sulfate
8240/8260
8270
8141, 8270, 8321
8270
8250/8270
8315
8250/8270
8270
8250/8270
8040, 8250/8270
8060/8061, 8250/8270, 8410
8090
8270
8270, 8330
8270
8250/8270, 8410
8040, 8250/8270, 8410
8090, 8250/8270, 8275, 8330,
8410
8090, 8250/8270, 8330, 8410
8270
8040, 8150/8151, 8270, 8321
8060/8061, 8250/8270, 8410
8410
8318
8240/8260
8141, 8270
8250/8270, 8275
8270
8250/8270
8321
8321
8321
8321
8321
8321
8321
8321
8321
8321
8140/8141, 8270, 8321
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
                                   TWO - 15
                   Revision 2
               September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound                                         Applicable Method(s)


Endrin                                           8080/8081,  8250/8270
Endrin aldehyde                                  8080/8081,  8250/8270
Endrin ketone                                    8081,  8250/8270
Epichlorohydrin                                  8010,  8240/8260
EPN                                              8141,  8270
Ethanol (Ethyl alcohol)                          8015,  8240/8260
Ethion                                           8141,  8270
Ethoprop                                         8140/8141
Ethyl acetate                                    8260
Ethyl benzene                                     8020,  8021,  8240/8260
Ethyl carbamate                                  8270
Ethylene dibromide                               8010,  8011,  8021,  8240/8260
Ethylene oxide                                   8240/8260
Ethyl methacrylate                               8240/8260
Ethyl methanesulfonate                           8250/8270
Ethyl parathion                                  8270
Etridiazole                                      8081
Famphur                                          8141,  8270,  8321
Fenitrothion                                     8141
Fensulfothion                                    8140/8141,  8270, 8321
Fenthion                                         8140/8141,  8270
Fluchloralin                                     8270
Fluoranthene                                     8100,  8250/8270, 8310,  8410
Fluorene                                         8100,  8250/8270, 8275,  8310,
                                                 8410
Fluorescent Brightener 61                        8321
Fluorescent Brightener 236                       8321
Fluorobenzene                                    8260
2-Fluorobiphenyl                                 8250/8270
2-Fluorophenol                                   8250/8270
Fonophos                                         8141
Formaldehyde                                     8315
Halowax-1000                                     8081
Halowax-1001                                     8081
Halowax-1013                                     8081
Hal owax-1014                                     8081
Halowax-1051                                     8081
Hal owax-1099                                     8081
Heptachlor                                       8080/8081,  8250/8270
Heptachlor epoxide                               8080/8081,  8250/8270
Heptanal                                         8315
Hexachlorobenzene                                8081,  8120/8121, 8250/8270,
                                                 8275,  8410
                                   TWO - 16                         Revision 2
                                                                September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
                                                 Applicable Method(s)
Hexachlorobutadiene (1,3-Hexachlorobutadiene)

Hexachlorocyclohexane
a-Hexachlorocyclohexane  (a-BHC)

0-Hexachl orocycl ohexane  (/3-BHC)

S-Hexachlorocyclohexane  (5-BHC)

7-Hexachlorocyclohexane  (7-BHC)

Hexachlorocyclopentadi ene

Hexachloroethane

Hexachlorophene
Hexachloropropene
Hexahydro-l,3,5-trinitro-l,3,5-triazine (RDX)
Hexamethylphosphoramide  (HMPA)
Hexanal
2-Hexanone
HMX
1,2,3,4,6,7,8-HpCDD
1,2,3,4,6,7,8-HpCDF
1,2,3,4,7,
1,2,3,4,7,
1,
1,
1.
L
L
          8,9-HpCDF
          8-HxCDD
  2,3,6,7,8-HxCDD
  2,3,7,8,9-HxCDD
  2,3,4,7,8-HxCDF
  2,3,6,7,8-HxCDF
  2,3,7,8,9-HxCDF
2,3,4,6,7,8-HxCDF
Hydroquinone
3-Hydroxycarbofuran
5-Hydroxydicamba
2-Hydroxypropi oni tri1e
Indeno(l,2,3-cd)pyrene
lodomethane
Isobutyl alcohol (2-Methyl-l-propanol)
Isodrin
Isophorone
Isopropylbenzene
p-Isopropyltoluene
Isosafrole
8021, 8120/8121, 8250/8270,
8260, 8410
8120
8080/8081, 8120/8121, 8250,
8270
8080/8081, 8120/8121, 8250,
8270
8080/8081, 8120/8121, 8250,
8270
8080/8081, 8120/8121, 8250,
8270
8081, 8120/8121, 8250/8270,
8410
8120/8121, 8250/8270, 8260,
8410
8270
8270
8330
8141, 8270
8315
8240/8260
8330
8280/8290
8280/8290
8280/8290
8280/8290
8280/8290
8280/8290
8280/8290
8280/8290
8280/8290
8280/8290
8270
8318
8151
8240/8260
8100, 8250/8270, 8310
8240/8260
8240/8260
8081, 8270
8090, 8250/8270, 8410
8021, 8260
8021, 8260
8270
                                   TWO - 17
                                                                    Revision 2
                                                                September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
Isovaleraldehyde
Kepone
Leptophos
Malathion
Maleic anhydride
Malononitrile
MCPA
MCPP
Merphos
Mestranol
Methacrylonitrile
Methanol
Methapyrilene
Methiocarb (Mesurol)
Methomyl (Lannate)
Methoxychlor (4,4'-Methoxychlor)
Methyl acrylate
Methyl-t-butyl ether
3-Methylcholanthrene
2-Methyl-4,6-dinitrophenol
4,4'-Methylenebis(2-chloroaniline)
4,4'-Methylenebis(N,N-dimethylaniline)
Methyl ethyl  ketone (MEK, 2-Butanone)
Methylene chloride (Dichloromethane)
Methyl iodide
Methyl isobutyl ketone (4-Methyl-2-pentanone)
Methyl methacrylate
Methyl methanesulfonate
2-Methylnaphthal ene
2-Methyl-5-nitroaniline
Methyl parathion
4-Methyl-2-pentanone (Methyl isobutyl ketone)
2-Methylphenol (o-Cresol)
3-Methylphenol (m-Cresol)
4-Methylphenol (p-Cresol)
2-Methylpyridine
Methyl-2,4,6-trinitrophenylnitramine  (Tetryl)
Mevinphos
Mexacarbate
Mi rex
Monochrotophos
Naled
Naphthalene
8315
8081, 8270
8141, 8270
8141, 8270
8270
8240/8260
8150/8151, 8321
8150/8151, 8321
8140/8141, 8321
8270
8240/8260
8260
8270
8318
8318, 8321
8080/8081, 8250/8270
8260
8260
8100, 8250/8270
8040
8270
8270
8015, 8240/8260
8010, 8021, 8240/8260
8010, 8240/8260
8015, 8240/8260
8240/8260
8250/8270
8250/8270, 8410
8270
8270, 8321
8015, 8240/8260
8250/8270, 8410
8270
8250/8270, 8275, 8410
8270
8330
8140/8141, 8270
8270
8081, 8270
8141, 8270, 8321
8140/8141, 8270, 8321
8021, 8100, 8250/8270, 8260,
8275, 8310, 8410
                                   TWO - 18
                    Revision  2
                September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
Naphthoquinone
1,4-Naphthoquinone
1-Naphthylamine
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine
Nitrobenzene  (NB)

4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
2-Nitropropane
Nitroquinoline-1-oxide
N-Ni trosodi butyl amine
N-Nitrosodiethyl amine
N-Nitrosodimethylamine
N-Ni trosodi phenylami ne
N-Ni trosodi-n-propylamine
N-Ni trosomethylethyl ami ne
N-Nitrosomorpholine
N-Nitrosopiperidine
N-Nitrosopyrrolidine
o-Nitrotoluene (2-NT)
m-Nitrotoluene (3-NT)
p-Nitrotoluene (4-NT)
5-Nitro-o-toluidine
trans-Nonachlor
Nonanal
OCDD
OCDF
Octahydro-l,3,5,7-tetranitro-l,3,5,7-tetrazocine
                           (HMX)
Octamethyl pyrophosphoramide
Octanal
4,4'-Oxydianiline
Parathion
Parathion, ethyl
Parathion, methyl
PCB-1016 (Aroclor-1016)
8090
8270
8250/8270
8250/8270
8270
8270
8250/8270, 8410
8250/8270, 8410
8250/8270, 8410
8270
8090, 8250/8270, 8260, 8330,
8410
8270
8081, 8270
8040, 8250/8270, 8410
8040, 8151, 8250/8270, 8410
8260
8270
8250/8270
8270
8070, 8250/8270, 8410
8070, 8250/8270, 8410
8070, 8250/8270, 8410
8270
8270
8250/8270
8270
8330
8330
8330
8270
8081
8315
8280/8290
8280/8290

8330
8270
8315
8270
8270
8141
8140/8141
8080/8081,  8250/8270
                                   TWO - 19
                   Revision 2
               September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
PCB-1221 (Aroclor-1221)
PCB-1232 (Aroclor-1232)
PCB-1242 (Aroclor-1242)
PCB-1248 (Aroclor-1248)
PCB-1254 (Aroclor-1254)
PCB-1260 (Aroclor-1260)
PCNB
1,2,3,4,7-PeCDD
1,2,3,7,8-PeCDD
1,2,3,7,8-PeCDF
2,3,4,7,8-PeCDF
Pentachlorobenzene
Pentachloroethane
Pentachlorohexane
Pentachloronitrobenzene
Pentachlorophenol
Pentaf1uorobenzene
Pentanal
trans-Permethrin
Perthane
Phenacetin
Phenanthrene

Phenobarbital
Phenol
1,4-Phenylenediamine
Phorate
Phosalone
Phosmet
Phosphamidion
Phthalic anhydride
Picloram
2-Picoline
Piperonyl sulfoxide
Promecarb
Pronamide
Propachlor
Propanal
Propargyl alcohol
6-Propiolactone
Propionitrile
Propoxur (Baygon)
n-Propylamine
n-Propylbenzene
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8080/8081, 8250/8270
8081
8280/8290
8280/8290
8280/8290
8280/8290
8121, 8250/8270
8240/8260
8120
8250/8270
8040, 8151, 8250/8270, 8410
8260, 4010
8315
8081
8081
8250/8270
8100, 8250/8270, 8275, 8310,
8410
8270
8040, 8250/8270, 8410
8270
8140/8141, 8270, 8321
8270
8141, 8270
8141, 8270
8270
8151
8240/8260, 8250/8270
8270
8318
8250/8270
8081
8315
8240/8260
8240/8260
8240/8260
8318
8240/8260
8021, 8260
                                   TWO  - 20
                    Revision 2
                September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
Propylthiouracil
Pyrene

Pyridine
RDX
Resorcinol
Ronnel
Safrole
Simazine
Solvent Red 3
Solvent Red 23
Stirophos (Tetrachlorvinphos)
Strobane
Strychnine
Styrene
Sulfall ate
Sulfotepp
2,4,5-T
2,4,5-T, butoxyethanol ester
2,4,5-T, butyl ester
1,2,3,4-TCDD
1,2,7,8-TCDD
1,2,8,9-TCDD
1,3,6,8-TCDD
1,3,7,8-TCDD
1,3,7,9-TCDD
2,3,7,8-TCDD
1,2,7,8-TCDF
2,3,7,8-TCDF
TEPP
Terbuphos (Terbufos)
Terphenyl
1,2,3,4-Tetrachlorobenzene
1,2,3,5-Tetrachlorobenzene
1,2,4,5-Tetrachlorobenzene
Tetrachlorobenzenes
1,1,1,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
2,3,4,6-Tetrachlorophenol
Tetrachlorophenols
Tetrachlorvinphos (Stirophos)
Tetraethyl dithiopyrophosphate
Tetraethyl pyrophosphate
8270
8100, 8250/8270, 8275, 8310,
8410
8240/8260, 8270
8330
8270
8140/8141
8270
8141
8321
8321
8140/8141, 8270
8081
8270, 8321
8021, 8240/8260
8270
8141
8150/8151, 8321
8321
8321
8280
8280
8280
8280
8280
8280
8280/8290
8280
8280/8290
8141
8141, 8270
8250/8270
8121
8121
8121, 8250/8270
8120
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8250/8270
8040
8140/8141, 8270
8270
8270
                                   TWO - 21
                   Revision 2
               September 1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound
Applicable Method(s)
Tetrazene
Thiofanox
Thionazine
Thiophenol (Benzenethiol)
TOCP (Tri-o-cresylphosphate)
Tokuthion (Prothiofos)
m-Tolualdehyde
o-Tolualdehyde
p-Tolualdehyde
Toluene
Toluene diisocyanate
o-Toluidine
Toxaphene
2,4,5-TP (Silvex)
2,4,6-Tribromophenol
1,2,3-Trichlorobenzene
1,2,4-Tri chlorobenzene

1,3,5-Tri chlorobenzene
1,1,1-Trichloroethane
1,1,2-Tri chloroethane
Trichloroethene
Trichlorofluoromethane
Trichlorfon
Trichloronate
2,4,5-Tri chlorophenol
2,4,6-Trichlorophenol
Trichlorophenols
1,2,3-Tri chloropropane
0,0,0-Triethyl phosphorothioate
Trifluralin
2,4,5-Trimethylaniline
1,2,4-Trimethylbenzene
1,3,5-Trimethylbenzene
Trimethyl phosphate
1,3,5-Trinitrobenzene (1,3,5-TNB)
2,4,6-Trinitrotoluene (2,4,6-TNT)
Tri-o-cresyl phosphate (TOCP)
Tri-p-tolyl phosphate
Tris(2,3-Dibromopropyl) phosphate (Tris-BP)
Vinyl acetate
Vinyl chloride
8331
8321
8141, 8270
8270
8141
8140/8141
8315
8315
8315
8020, 8021, 8240/8260
8270
8270
8080/8081, 8250/8270
8150/8151, 8321
8250/8270
8021, 8121, 8260
8021, 8120/8121, 8250/8270,
8260, 8410
8121
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8010, 8021, 8240/8260
8141, 8321
8140/8141
8250/8270, 8410
8040, 8250/8270, 8410
8040
8010, 8021, 8240/8260
8270
8081, 8270
8270
8021, 8260
8021, 8260
8270
8270, 8330
8330
8141
8270
8270, 8321
8240/8260
8010, 8021, 8240/8260
                                   TWO - 22
                    Revision  2
                September  1994

-------
                                  TABLE 2-1.
                                  (Continued)
Compound                                         Applicable Method(s)
o-Xylene                                         8021, 8260
m-Xylene                                         8021, 8260
p-Xylene                                         8021, 8260
Xylene (Total)                                   8020, 8240
                                  TABLE 2-2A.
                      METHOD 3650 -  BASE/NEUTRAL FRACTION


Benz(a)anthracene                               Hexachlorobenzene
Benzo(a)pyrene                                  Hexachlorobutadiene
Benzo(b)fluoranthene                            Hexachloroethane
Chi ordane                                       Hexachlorocyclopentadi ene
Chlorinated dibenzodioxlns                      Naphthalene
Chrysene                                        Nitrobenzene
Creosote                                        Phorate
Dichlorobenzene(s)                              2-Picoline
Dinitrobenzene                                  Pyridine
2,4-Dinitrotoluene                              Tetrachlorobenzene(s)
Heptachlor                                      Toxaphene
                                  TABLE  2-2B.
                          METHOD  3650  -  ACID  FRACTION

2-Chlorophenol                                  4-Nitrophenol
Cresol(s)                                       Pentachlorophenol
Creosote                                        Phenol
Dichlorophenoxyacetic acid                      Tetrachlorophenol(s)
2,4-Dimethyl phenol                              Trichlorophenol(s)
4,6-Dinitro-o-cresol                            2,4,5-TP (Silvex)
                                   TWO - 23                         Revision 2
                                                                September 1994

-------
                                  TABLE 2-3.
                    METHOD 5041 - SORBENT CARTRIDGES FROM
                    VOLATILE ORGANIC SAMPLING TRAIN (VOST)

Acetone                                         1,2-Dichloropropane
Acrylonitrlle                                   cis-l,3-Dichloropropene
Benzene                                         trans-l,3-Dichloropropene
Bromodichloromethane                            Ethyl benzene3
Bromoform8                                      lodomethane
Bromomethane                                    Methylene chloride
Carbon disulfide                                Styrene3
Carbon tetrachloride                            1,1,2,2-Tetrachloroethane3
Chlorobenzene                                   Tetrachloroethene
Chlorodibrompmethane                            Toluene
Chi oroethane                                    1,1,1-Tri chloroethane
Chloroform                                      1,1,2-Trichloroethane
Chloromethane                                   Trichloroethene
Di bromomethane                                  Tri chlorof1uoromethane
1,1-Di chloroethane                              1,2,3-Tri chloropropane3
1,2-Dichloroethane                              Vinyl  chloride
1,1-Dichloroethene                              Xylenes3
trans-l,2-Dichloroethene


3  Boiling point of this compound is above 132°C.   Method 0030  is  not
appropriate for quantitative sampling of this analyte.

b  Boiling point of this compound is below 30°C.   Special  precautions  must be
taken when sampling for this analyte by Method 0030.  Refer to Sec. 1.3 of
Method 5041 for discussion.
                                   TWO - 24                         Revision 2
                                                                September 1994

-------
                                  TABLE 2-4.
                      METHOD 8010 -  HALOGENATED VOLATILES
AHyl chloride
Benzyl chloride
Bi s(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Bromoacetone
Bromobenzene
Bromodichloromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethanol
2-Chloroethyl vinyl ether
Chloroform
1-Chlorohexane
Chloromethane
Chloromethyl methyl ether
Chloroprene
4-Chlorotoluene
Di bromochloromethane
l,2-Dibromo-3-chloropropane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Di chlorobenzene
l,4-Dichloro-2-butene
Di chlorodi f1uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene (Vinylidene chloride)
trans-1,2-Dichloroethene
Dichloromethane (Methylene Chloride)
1,2-Di chloropropane
1,3-Di chloro-2-propanol
cis-l,3-Dichloropropene
trans-1,3-Di chloropropene
Epichlorhydrin
Ethylene dibromide
Methyl iodide
1,1,2,2-Tetrachloroethane
1,1,1,2-Tetrachloroethane
Tetrachloroethene
1,1,1-Trichloroethane
1,1,2-Tri chloroethane
Trichloroethene
Tri chlorof1uoromethane
1,2,3-Tri chloropropane
Vinyl chloride
                    For Method 8011,  see Table 2-7
            TABLE 2-5.
METHOD 8015 - NONHALOGENATED VOLATILES
                   TABLE  2-6.
           METHOD 8020 -  AROMATIC VOLATILES
Di ethyl ether
Ethanol
Methyl ethyl ketone (MEK)
Methyl isobutyl ketone  (MIBK)
           Benzene
           Chlorobenzene
           1,2-Dichlorobenzene
           1,3-Dichlorobenzene
           1,4-Dichlorobenzene
           Ethyl benzene
           Toluene
           Xylenes
                                   TWO - 25
                                  Revision  2
                              September  1994

-------
                                        TABLE 2-7.
              METHOD 8021  (METHOD 8011*)  -  HAL06ENATED AND AROMATIC VOLATILES

Benzene                                         1,3-Dichloropropane
Bromobenzene                                    2,2-Dichloropropane
Bromochloromethane                              1,1-Di chloropropene
Bromodi chloromethane                            ci s-1,3-Di chloropropene
Bromoform                                       trans-1,3-Di chloropropene
Bromomethane                                    Ethyl benzene
n-Butylbenzene                                  Hexachlorobutadi ene
sec-Butyl benzene                                Isopropylbenzene
tert-Butylbenzene                               p-Isopropyltoluene
Carbon tetrachloride                            Methylene chloride  (DCM)
Chlorobenzene                                   Naphthalene
Chlorodibromomethane                            n-Propylbenzene
Chloroethane                                    Styrene
Chloroform                                      1,1,1,2-Tetrachloroethane
Chloromethane                                   1,1,2,2-Tetrachloroethane
2-Chlorotoluene                                 Tetrachloroethene
4-Chlorotoluene                                 Toluene
l,2-Dibromo-3-chloropropane*                    1,2,3-Trichlorobenzene
1,2-Dibromoethane*                              1,2,4-Trichlorobenzene
Di bromomethane                                  1,1,1-Tri chloroethane
1,2-Di chlorobenzene                             1,1,2-Tri chloroethane
1,3-Di chlorobenzene                             Tri chloroethene
1,4-Di chlorobenzene                             Tri chlorof1uoromethane
Di chlorodi f1uoromethane                         1,2,3-Tri chloropropane
1,1-Dichloroethane                              1,2,4-Trimethylbenzene
1,2-Di chloroethane                              1,3,5-Tri methyl benzene
1,1-Dichloroethene  (Vinylidene chloride)        Vinyl chloride
cis-l,2-Dichloroethene                          o-Xylene
trans-1,2-Di chloroethene                        m-Xylene
1,2-Dichloropropane                             p-Xylene

      *  Indicates the only two target analytes of Method 8011.  These constituents are
                                 also target analytes of Method 8021.
           TABLE 2-8.                                           TABLE 2-9
      METHODS 8030/8031 -                                     METHOD 8032 -
      ACROLEIN, ACRYLONITRILE                                   ACRYLAMIDE

      Acrolein (Propenal)*                                  Acrylamide
      Acrylonitrile

  *  Target analyte of Method 8030 only.
                                         TWO - 26                         Revision 2
                                                                      September 1994

-------
                                        TABLE 2-10.
                                  METHOD 8040 - PHENOLS
2-sec-Butyl-4,6-dinitrophenol (DNBP, Dinoseb)
4-Chloro-3-methylphenol
2-Chlorophenol
Cresols (Methylphenols)
2-Cyclohexyl-4,6-di ni trophenol
2,4-Dichlorophenol
2,6-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
    2-Methyl-4,6-dinitrophenol
    2-Nitrophenol
    4-Nitrophenol
    Pentachlorophenol
    Phenol
    Tetrachlorophenols
    2,4,6-Trichlorophenol
    Trichlorophenols
            TABLE 2-11.
   METHODS 8060/8061 - PHTHALATE ESTERS

Benzyl benzoate*
Butyl benzyl phthalate
Bis(2-ethylhexyl) phthalate
Di-n-butyl phthalate
Diethyl phthalate
Dimethyl phthalate
Di-n-octyl phthalate

*  Target analyte of Method 8061 only.
       TABLE 2-12.
METHOD 8070 - NITROSAMINES

    N-Nitrosodimethylamine
    N-Nitrosodiphenylamine
    N-Nitrosodi-n-propylamine
                                         TWO - 27
                    Revision 2
                September 1994

-------
                                  TABLE 2-13.
            METHODS 8080/8081 - ORGANOCHLORINE PESTICIDES AND PCBs
Aroclor-1016  (PCB-1016)
Aroclor-1221  (PCB-1221)
Aroclor-1232  (PCB-1232)
Aroclor-1242  (PCB-1242)
Aroclor-1248  (PCB-1248)
Aroclor-1254  (PCB-1254)
Aroclor-1260  (PCB-1260)
Alachlor*
Aldrin
a-BHC
0-BHC
5-BHC
7-BHC (Lindane)
Captafol*
Captan*
Chiorobenzilate*
Chlordane (technical)**
a-Chlordane*
y-Chlordane*
Chloroneb*
Chloropropylate*
Chlorothalonil*
DBCP*
DC PA*
4,4'-DDD
4,4'-DDE
4,4'-DDT
Diallate*
Dichlone*
Dicofol*
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone*
Etridiazole*
Halowax-1000*
Halowax-1001*
Halowax-1013*
*  Target analyte of Method 8081 only.
** Target analyte of Method 8080 only.
Halowax-1014*
Halowax-1051*
Halowax-1099*
Heptachlor
Heptachlor epoxide
Hexachlorobenzene*
Hexachlorocyclo-
  pentadiene*
Isodrin*
Kepone*
Methoxychlor
Mi rex*
Nitrofen*
trans-Nonachlor*
PCNB*
trans-Permethrin*
Perthane*
Propachlor*
Strobane*
Toxaphene
Trifluralin*
                               TABLE 2-14.
                        METHOD 8090 - NITROAROMATICS AND
                              CYCLIC KETONES

                              Dinitrobenzene
                              2,4-Dinitrotoluene
                              2,6-Dinitrotoluene
                              Isophorone
                              Naphthoquinone
                              Nitrobenzene
                                   TWO - 28
                                          Revision 2
                                      September 1994

-------
                                      TABLE 2-15.
                   METHODS 8100 - POLYNUCLEAR AROMATIC HYDROCARBONS
    Acenaphthene
    Acenaphthylene
    Anthracene
    Benzo(a)anthracene
    Benzo(b)fl uoranthene
    Benzo(j)f 1uoranthene
    Benzo(k)f1uoranthene
    Benzo (g,h,i)perylene
    Benzo(a)pyrene
    Chrysene
    Dibenz(a,h)acridine
    Dibenz(a,j)acridine
    D1benzo(a,h)anthracene
        7H-Dibenzo(c,g)carbazole
        Dibenzo(a,e)pyrene
        Dibenzo(a,h)pyrene
        Dibenzo(a,i)pyrene
        Fluoranthene
        Fluorene
        Indeno(l,2,3-cd)pyrene
        3-Methylcholanthrene
        Naphthalene
        Phenanthrene
        Pyrene
                                     TABLE 2-16
                               METHOD 8110 -  HALOETHERS
          Bi s(2-Chloroethoxy)methane
          Bis(2-Chloroethyl) ether
          Bis(2-Chloroisopropyl) ether
        4-Bromophenyl phenyl ether
        4-Chlorophenyl phenyl ether
                                      TABLE 2-17.
                     METHODS 8120/8121 - CHLORINATED HYDROCARBONS
Benzal chloride*
Benzotrichloride*
Benzyl chloride*
2-Chloronaphthalene
1,2-Di chlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobutadi ene
Hexachlorocyclohexane**
a-Hexachlorocyclohexane (a-BHC)*
)S-Hexachl orocycl ohexane (/3-BHC)*
6-Hexachlorocyclohexane (6-BHC)*
Y-Hexachlorocyclohexane (y-BHC)*
Hexachlorocyclopentadi ene
Hexachloroethane
Pentachlorobenzene*
Pentachlorohexane**
Tetrachlorobenzenes**
1,2,3,4-Tetrachlorobenzene*
1,2,3,5-Tetrachlorobenzene*
1,2,4,5-Tetrachlorobenzene*
1,2,3-Tri chlorobenzene*
1,2,4-Tri chlorobenzene
1,3,5-Tri chlorobenzene*
*  Target analyte of Method 8121 only.
** Target analyte of Method 8120 only.
                                       TWO - 29
                            Revision 2
                        September 1994

-------
                                  TABLE  2-18.
                METHODS 8140/8141 - ORGANOPHOSPHORUS COMPOUNDS
                        (PACKED AND CAPILLARY COLUMNS)


 Aspon*                              Fenthion
 Atrazine*                           Fonophos*
 Azinphos ethyl*                     Hexamethylphosphoramide*  (HMPA)
 Azinphos methyl                     Leptophos*
 Bolstar (Sulprofos)                 Malathion*
 Carbophenothion*                    Merphos
 Chlorofenvinphos*                   Mevinphos
 Chlorpyrifos                        Monochrotophos*
 Chlorpyrifos methyl*                Naled
 Coumaphos                           Parathion,  ethyl*
 Crotoxypos*                         Parathion,  methyl
 Demeton-0, and -S                   Phorate
 Diazinon                            Phosmet*
 Dichlorofenthion*                   Phosphamidon*
 Dichlorvos (DDVP)                   Ronnel
 Dichrotophos*                       Simazine*
 Dimethoate*                         Stirophos  (Tetrachlorvinphos)
 Dioxathion*                         Sulfotep*
 Disulfoton                          TEPP*
 EPN*                                Terbufos*
 Ethion*                             Thionazin*
 Ethoprop                            Tokuthion  (Prothiofos)
 Famphur*                            Trichlorfon*
 Fenitrothion*                       Trichloronate
 Fensulfothion                       Tri-o-cresylphosphate  (TOCP)*
      *  Target analyte of Method 8141 only.
                                  TABLE 2-19.
                  METHODS 8150/8151 - CHLORINATED HERBICIDES


Acifluorfen*             Dicamba                      MCPA
Bentazon*                3,5-Dichlorobenzoic acid*    MCPP
Chloramben*              Dichlorprop                  4-Nitrophenol*
2,4-D                    Dinoseb  (DNBP)               Pentachlorophenol*
Dalapon                  5-Hydroxydicamba*            Picloram*
2,4-DB                                                2,4,5-TP (Silvex)
DCPA diacid*                                          2,4,5-T

   *  Target analyte of Method 8151 only.
                                   TWO - 30                         Revision 2
                                                                September 1994

-------
Acetone
Acetonitrile
Acrolein (Propenal)
Acrylonitrile
Ally! alcohol
Allyl chloride
Benzene
Benzyl chloride
Bis(Z-chloroethyl) sulfide
Bromoacetone
Bromobenzene*
Bromochloromethane
Bromodichloromethane
4-Bromof1uorobenzene
Bromoform
Bromomethane
n-Butanol*
2-Butanone (Methyl ethyl
ketone)
n-Butylbenzene*
sec-Butyl benzene*
tert-Butylbenzene*
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chioroaceton i tri1e*
Chlorobenzene
2-Chloro-l,3-butadiene*
1-Chlorobutane*
Chiorodi bromomethane
Chloroethane
2-Chloroethanol
2-Chloroethyl vinyl ether
Chloroform
1-Chlorohexane*
Chioromethane
Chloroprene
3-Chloropropene*
3-Chioropropi oni tri1e
2-Chlorotoluene*
4-Chlorotoluene*
Crotonaldehyde*
                                        TABLE 2-20.
                               METHODS 8240/8260 - VOLATILES
l,2-Dibromo-3-
 chloropropane
1,2-Dibromoethane
Dibromomethane
Di bromof1uoromethane*
1,2-Di chlorobenzene*
1,3-Dichlorobenzene*
1,4-Di chlorobenzene*
l,4-Dichloro-2-butene**
cis-l,4-Dichloro-
 2-butene*
trans-l,4-Dichloro-2-
  butene*
l,4-Dichloro-2-butene**
Di chlorodi f1uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
ci s-1,2-Di chloroethene*
trans-1,2-Dichloroethene
1,2-Dichloropropane
1,3-Dichloropropane*
2,2-Di chloropropane*
l,3-Dichloro-2-propanol
1,1-Di chloropropene*
cis-l,3-Dichloropropene
trans-1,3-Dichloropropene
1,2,3,4-Diepoxybutane
Diethyl  ether*
1,4-Di f1uorobenzene
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl acetate*
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
Fluorobenzene*
Hexachlorobutadiene*
Hexachloroethane*
2-Hexanone
2-Hydroxypropionitrile
lodotnethane
Isobutyl alcohol
Isopropylbenzene*
p-Isopropyltoluene*
Malononitrile
Methacrylonitrile
Methanol*
Methyl acrylate*
Methyl-t-butyl ether*
Methylene chloride (DCM)
Methyl iodide
Methyl methacrylate
4-Methyl-2-pentanone
  (MIBK)
Naphthalene*
Nitrobenzene*
2-Nitropropane*
Pentachloroethane
Pentaf1uorobenzene*
2-Picoline
Propargyl alcohol
B-Propiolactone
Propionitrile
n-Propylamine
n-Propylbenzene*
Pyridine
Styrene
1,1,1,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,2,3-Tri chlorobenzene*
1,2,4-Tri chlorobenzene*
1,1,1-Tri chloroethane
1,1,2-Tri chloroethane
Trichloroethene
Tri chlorof1uoromethane
1,2,3-Trichloropropane
1,2,4-Trimethylbenzene*
1,3,5-Trimethylbenzene*
Vinyl acetate
Vinyl chloride
Xylene (Total)**
o-Xylene*
m-Xylene*
p-Xylene*
* Target analyte of Method 8260 only.
** Target analyte of Method 8240 only.
                                         TWO - 31
                                          Revision 2
                                      September 1994

-------
                                        TABLE 2-21.
                             METHODS 8250/8270 -  SEMIVOLATILES
Acenaphthene
Acenaphthylene
Acetophenone
2-Acetylami nof1uorene*
1-Acetyl-2-thi ourea*
Aldrin
2-Aminoanthraquinone*
Aminoazobenzene*
4-Aminobiphenyl
3-Amino-9-ethylcarbazole*
Anilazine*
Aniline
o-Anisidine*
Anthracene
Aramite*
Aroclor-1016 (PCB-1016)
Aroclor-1221 (PCB-1221)
Aroclor-1232 (PCB-1232)
Aroclor-1242 (PCB-1242)
Aroclor-1248 (PCB-1248)
Aroclor-1254 (PCB-1254)
Aroclor-1260 (PCB-1268)
Azinphos-methyl*
Barban*
Benz(a)anthracene
Benzidine
Benzo(b)f1uoranthene
Benzo(k)f1uoranthene
Benzoic acid
Benzo(g,h,i)perylene
Benzo(a)pyrene
p-Benzoquinone*
Benzyl alcohol
a-BHC
)8-BHC
5-BHC
•y-BHC (Lindane)
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Bromoxynil*
Butyl benzyl phthalate
2-sec-Butyl-4,6-dinitrophenol  (Dinoseb)*
Captafol*
Captan*
Carbaryl*
Carbofuran*
Carbophenothion*
Chlordane (technical)
Chlorfenvinphos*
4-Chloroaniline
Chiorobenzilate*
5-Chloro-2-methylaniline*
4-Chloro-3-methyl phenol
3-(Chloromethyl)pyridine hydrochloride*
1-Chloronaphthalene
2-Chloronaphthalene
2-Chlorophenol
4-Chloro-l,2-phenylenediamine*
4-Chloro-l,3-phenylenediamine*
4-Chlorophenyl phenyl ether
Chrysene
Coumaphos*
p-Cresidine*
Crotoxyphos*
2-Cyclohexyl-4,6-dinitrophenol*
4,4'-DDD
4,4'-DDE*
4,4'-DDT
Demeton-0*
Demeton-S*
Diallate (cis or trans)*
2,4-Diaminotoluene*
Dibenz(a,j)acridine
Dibenz(a,h)anthracene
Dibenzofuran
Di benzo(a,e)pyrene*
l,2-Dibromo-3-chloropropane*
Di-n-butyl phthalate
Dichlone*
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos*
Dicrotophos*
Dieldrin
Diethyl phthalate
Di ethylsti1bestrol*
Diethyl sulfate*
Dihydrosaffrole*
Dimethoate*
3,3'-Dimethoxybenzidine*
Dimethyl aminoazobenzene
                                         TWO - 32
                          Revision 2
                      September 1994

-------
                                        TABLE 2-21.
                       METHODS 8250/8270 -  SEMIVOLATILES  (CONTINUED)
7,12-Dimethylbenz(a)anthracene
3,3'-Dimethylbenzidine*
a,a-Dimethylphenethylamine
2,4-Dimethylphenol
Dimethyl phthalate
1,2-Dinitrobenzene*
1,3-Dinitrobenzene*
1,4-Dinitrobenzene*
4,6-Dinitro-2-methylphenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap*
Dioxathion*
Diphenylamine
5,5-Di phenylhydantoi n*
1,2-Diphenylhydrazine
Di-n-octyl phthalate
Disulfoton*
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
EPN*
Ethion*
Ethyl carbamate*
Ethyl methanesulfonate
Ethyl parathion*
Famphur*
Fensulfothion*
Fenthion*
Fluchloralin*
Fluoranthene
Fluorene
2-Fluorobiphenyl
2-Fluorophenol
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadi ene
Hexachlorocyclopentadi ene
Hexachloroethane
Hexachlorophene*
Hexachloropropene*
Hexamethyl phosphoratnide*
Hydroquinone*
Indeno(l,2,3-cd)pyrene
Isodrin*
Isophorone
Isosafrole*
Kepone*
Leptophos*
Malathion*
Maleic anhydride*
Mestranol*
Methapyrilene*
Methoxychlor
3-Methylcholanthrene
4,4'-Methylenebis(2-chloroaniline)*
4,4'-Methylenebis(N,N-dimethylaniline)*
Methyl methanesulfonate
2-Methylnaphtha!ene
2-Methyl-5-nitroaniline*
Methyl parathion*
2-Methylphenol (o-Cresol)
3-Methylphenol (m-Cresol)*
4-Methylphenol (p-Cresol)
2-Methylpyridine*
Mevinphos*
Mexacarbate*
Mi rex*
Monocrotophos*
Naled*
Naphthalene
1,4-Naphthoquinone*
1-Naphthylamine
2-Naphthylamine
Nicotine*
5-Nitroacenaphthene*
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine*
Nitrobenzene
4-Nitrobiphenyl*
Nitrofen*
2-Nitrophenol
4-Nitrophenol
Nitroquinoline-1-oxide*
N-Nitrosodibutyl amine
N-Nitrosodiethyl amine*
N-Nitrosodimethylamine
N-Nitrosodiphenylamine
N-Nitrosodi-n-propylamine
                                         TWO - 33
                          Revision 2
                      September 1994

-------
                                        TABLE 2-21.
                       METHODS 8250/8270 -  SEMIVOLATILES (CONTINUED)
N-Nitrosomethylethylamine*
N-Nitrosomorpholine*
N-Nitrosopiperi dine
N-Nitrosopyrrolidine*
5-Nitro-o-toluidine*
Octamethyl pyrophosphoramide*
4,4'-Oxydianiline*
Parathion*
Pentachlorobenzene
Pentachloroni trobenzene
Pentachlorophenol
Phenacetin
Phenanthrene
Phenobarbital*
Phenol
1,4-Phenylenediamine*
Phorate*
Phosalone*
Phosmet*
Phosphamidion*
Phthalic anhydride*
2-Picoline
Piperonyl sulfoxide*
Pronamide
Propylthiouracil*
'yrene
}yr idine*
Resorcinol*
Safrole*
Strychnine*
Sulfall ate*

*  Target analyte of Method 8270 only.
                 Terbuphos*
                 Terphenyl
                 1,2,4,5-Tetrachlorobenzene
                 2,3,4,6-Tetrachlorophenol
                 Tetrachlorvinphos  (Stirophos)*
                 Tetraethyl  dithiopyrophosphate*
                 Tetraethyl  pyrophosphate*
                 Thionazine*
                 Thiophenol  (Benzenethiol)*
                 Toluene diisocyanate*
                 o-Toluidine*
                 Toxaphene
                 2,4,6-Tribromophenol
                 1,2,4-Tri chlorobenzene
                 2,4,5-Trichlorophenol
                 2,4,6-Trichlorophenol
                 0,0,0-Triethyl  phosphorothioate*
                 Trifluralin*
                 2,4,5-Trimethylaniline*
                 Trimethyl phosphate*
                 1,3,5-Tri n i trobenzene*
                 Tris(2,3-dibromopropyl)  phosphate*
                 Tri-p-tolyl phosphate*
                                        TABLE 2-22.
                          METHOD 8275 -  SEMIVOLATILES (SCREENING)
Aldrin
Benzo(k)fluoranthene
Benzo(a)pyrene
Carbazole
4-Chloro-3-methylphenol
1-Chloronaphthalene
2-Chlorophenol
Dibenzothiophene
2,4-Dichlorophenol
2,4-Dinitrotoluene
Diphenylamine
Fluorene
Hexachlorobenzene
4-Methylphenol
Naphthalene
Phenanthrene
Pyrene
                                         TWO - 34
                                            Revision  2
                                        September  1994

-------
2,3,7,8-TCDD
1,2,3,4-TCDD*
1,3,6,8-TCDD*
1,3,7,9-TCDD*
1,3,7,8-TCDD*
1,2,7,8-TCDD*
1,2,8,9-TCDD*
                               TABLE  2-23.
               METHODS  8280/8290  -  DIOXINS AND DIBENZOFURANS
1,2,3,4,7-PeCDD
1,2,3,7,8-PeCDD
1,2,3,4,7,8-HxCDD
1,2,3,6,7,8-HxCDD
1,2,3,7,8,9-HxCDD
1,2,3,4,6,7,8-HpCDD
OCDD
    * Target  analyte  of 8280  only
1,2,7,8-TCDF
2,3,7,8-TCDF
1,2,3,7,8-PeCDF
2,3,4,7,8-PeCDF
1,2,3,4,7,8-HxCDF
1,2,3,6,7,8-HxCDF
1,2,3,7,8,9-HxCDF
2,3,4,6,7,8-HxCDF
1,2,3,4,6,7,8-HpCDF
1,2,3,4,7,8,9-HpCDF
OCDF
                               TABLE  2-24.
              METHOD  8310  -  POLYNUCLEAR AROMATIC HYDROCARBONS
              Acenaphthene
              Acenaphthylene
              Anthracene
              Benzo(a)anthracene
              Benzo(a)pyrene
              Benzo (b)fl uoranthene
              Benzo(g,h,i)perylene
              Benzo(k)f1uoranthene
                     Chrysene
                     Di benzo(a,h)anthracene
                     Fluoranthene
                     Fluorene
                     Indeno(l,2,3-cd)pyrene
                     Naphthalene
                     Phenanthrene
                     Pyrene
                                 TWO -  35
                                         Revision  2
                                     September 1994

-------
                           TABLE 2-25.
                 METHOD 8315  -  CARBONYL  COMPOUNDS

   Acetaldehyde                  Heptanal
   Acetone                       Hexanal (Hexaldehyde)
   Acrolein (Propanol)           Isovaleraldehyde
   Benzaldehyde                  Nonanal
   Butanal (Butyraldehyde)       Octanal
   Crotonaldehyde                Pentanal (Valeraldehyde)
   Cyclohexanone                 Propanal (Propionaldehyde)
   Decanal                       m-Tolualdehyde
   2,5-Dimethylbenzaldehyde      o-Tolualdehyde
   Formaldehyde                  p-Tolualdehyde
       TABLE 2-26.                                  TABLE 2-27.
METHOD 8316 - ACRYLAMIDE,                 METHOD 8318 -  N-METHYLCARBAMATES
ACRYLONITRILE AND ACROLEIN
                                                Aldicarb  (Temik)
       Acrolein (Propanol)                      Aldicarb  Sulfone
       Acrylamide                               Carbaryl  (Sevin)
       Acrylonitrile                            Carbofuran (Furadan)
                                                Dioxacarb
                                                3-Hydroxycarbofuran
                                                Methiocarb (Mesurol)
                                                Methomyl  (Lannate)
                                                Promecarb
                                                Propoxur  (Baygon)
                             TWO - 36                         Revision 2
                                                          September 1994

-------
                                        TABLE 2-28.
                                METHOD 8321  - NONVOLATILES
    Azo Dves
Disperse Red 1
Disperse Red 5
Disperse Red 13
Disperse Yellow 5
Disperse Orange 3
Disperse Orange 30
Disperse Brown 1
Solvent Red 3
Solvent Red 23

Anthraquinone Dves
Disperse Blue 3
Disperse Blue 14
Disperse Red 60
Coumarin Dyes

(Fluorescent Brighteners)
Fluorescent Brightener 61
Fluorescent Brightener 236

Chlorinated Phenoxvacid Compounds
2,4-D
2,4-D, butoxyethanol ester
2,4-D, ethylhexyl ester
2,4-DB
Dalapon
Dicamba
Dichlorprop
Dinoseb
MCPA
MCPP
Silvex (2,4,5-TP)
2,4,5-T
2,4,5-T, butyl  ester
2,4,5-T, butoxyethanol ester
Alkaloids
Strychnine
Orqanophosphorus Compounds
Asulam
Dichlorvos
Dimethoate
Disulfoton
Famphur
Fensulfothion
Merphos
Methomyl
Methyl parathion
Monocrotophos
Naled
Phorate
Trichlorfon
Thiofanox
Tris-(2,3-dibromopropyl) phosphate,
  (Tris-BP)
                                         TWO - 37
                    Revision 2
                September 1994

-------
                           TABLE 2-29.
           METHOD 8330 - NITROAROMATICS AND NITRAMINES

4-Amino-2,6-dinitrotoluene (4-Am-DNT)
2-Amino-4,6-dinitrotoluene (2-Am-DNT)
1,3-Dinitrobenzene (1,3-DNB)
2,4-Dinitrotoluene (2,4-DNT)
2,6-Dinitrotoluene (2,6-DNT)
Hexahydro-l,3,5-trinitro-l,3,5-triazine (RDX)
Methyl-2,4,6-tri ni trophenylnitramine (Tetryl)
Nitrobenzene (NB)
2-Nitrotoluene (2-NT)
3-Nitrotoluene (3-NT)
4-Nitrotoluene (4-NT)
Octahydro-l,3,5,7-tetranitro-l,3,5,7-tetrazocine (HMX)
1,3,5-Trinitrobenzene (1,3,5-TNB)
2,4,6-Trinitrotoluene (2,4,6-TNT)
                           TABLE 2-30.
                     METHOD 8331 - TETRAZENE

                             Tetrazene
                             TWO - 38                         Revision 2
                                                          September 1994

-------
                              TABLE 2-31
                      METHOD 8410  -  SEMIVOLATILES
Acenaphthene
Acenaphthylene
Anthracene
Benzo(a)anthracene
Benzo(b)pyrene
Benzole acid
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl)ether
Bis(2-chloroisopropyl)ether
Bi s(2-ethylhexyl)phthalate
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
4-Chloroaniline
4-Chloro-3-methylphenol
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Dibenzofuran
Di-n-butyl phthalate
1,2-Di chlorobenzene
1,3-Di chlorobenzene
1,4-Di chlorobenzene
2,4-Dichlorophenol
Diethyl phthalate
Dimethyl phthalate
4,6-Dinitro-2-methylphenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Di-n-propyl phthalate
Fluoranthene
Fluorene
Hexachlorobenzene
1,3-Hexachlorobutadiene
Hexachlorocycl opentadi ene
Hexachloroethane
Isophorone
2-Methylnaphthalene
2-Methylphenol
4-Methylphenol
Naphthalene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
2-Nitrophenol
4-Nitrophenol
N-Nitrosodimethylamine
N-Nitrosodiphenylamine
N-Nitroso-di-n-propylamine
Pentachlorophenol
Phenanthrene
Phenol
Pyrene
1,2,4-Trichlorobenzene
2,4,5-Tri chlorophenol
2,4,6-Trichlorophenol
                               TWO - 39
                              Revision 2
                          September 1994

-------
               TABLE  2-32.
ANALYSIS METHODS FOR INORGANIC COMPOUNDS
Compound
Aluminum
Antimony
Arsenic
Barium
Beryl 1 i urn
Bromide
Cadmium
Calcium
Chloride
Chromium
Chromium, hexavalent
Cobalt
Copper
Cyanide
Fluoride
Iron
Lead
Lithium
Magnesium
Manganese
Mercury
Molybdenum
Nickel
Nitrate
Nitrite
Osmium
Phosphate
Phosphorus
Potassium
Selenium
Silver
Sodium
Strontium
Sulfate
Sulfide
Thallium
Tin
Vanadium
Zinc
Applicable Method (s)
6010,
6010,
6010,
6010,
6010,
9056
6010,
6010,
9056,
6010,
7195,
6010,
6010,
9010,
9056
6010,
6010,
6010,
6010,
6010,
7470,
6010,
6010,
9056,
9056
7550
9056
6010
6010,
6010,
6010,
6010,
6010,
9035,
9030,
6010,
7870
6010,
6010,
6020,
6020,
6020,
6020,
6020,

6020,
7140
9250,
6020,
7196,
6020,
6020,
9012,

7380,
6020,
7430
7450
6020,
7471
7480,
6020,
9200




7610
7740,
6020,
7770
7780
9036,
9031
6020,

7910,
6020,
7020
7040,
7060,
7080,
7090,

7130,

9251,
7190,
7197,
7200,
7210,
9013

7381
7420,


7460,

7481
7520






7741,
7760,


9038,

7840,

7911
7950,

7041,
7061,
7081
7091

7131

9252,
7191
7198
7201
7211



7421


7461









7742
7761


9056

7841


7951

7062
7062





9253






























                TWO - 40
    Revision 2
September 1994

-------
                                                        TABLE 2-33.
                        CONTAINERS, PRESERVATION TECHNIQUES,  AND  HOLDING TIMES FOR AQUEOUS MATRICESA
    Name
                                       Container
                                                         Preservation
                                                                                    Maximum holding time
    Bacterial Tests:
      Coliform, total                  P, G
    Inorganic Tests:
      Chloride                         P, G
      Cyanide, total and amenable      P, G
       to chlorination
                                                       Cool,  4°C, 0.008% Na2S203

                                                       None required
                                                       Cool,  4°C; if oxidizing
                                                       agents present  add 5 mL
                                                       0.1N NaAs02 per L or 0.06 g
                                                       of ascorbic  acid per L;
                                                       adjust pH>12 with 50% NaOH.
                                                       See Method 9010 for other
                                                       interferences.
6 hours

28 days
14 days
Hydrogen ion (pH)
Nitrate
Sulfate
Sulfide
Metals:
Chromium VI
Mercury
Metals, except chromium VI
and mercury
Organic Tests:
Acrolein and acrylonitri le

Benzidines


Chlorinated hydrocarbons

Dioxins and Furans

Haloethers

Nitroaromatics and
cyclic ketones
Nitrosamines

Oil and grease
Organic carbon, total (TOO
PCBs

Pesticides

Phenols

Phthalate esters

Polynuclear aromatic
hydrocarbons
Purgeable aromatic
hydrocarbons
Purgeable Halocarbons

Total organic ha I ides (TOX)

Radiological Tests:
Alpha, beta and radium
P, G
P, G
P, G
P, G

P, G
P, G
P, G


G, Teflon- lined
septum
G, Teflon-lined
cap

G, Teflon- lined
cap
G, Teflon- lined
cap
G, Teflon- lined
cap
G, Teflon-lined
cap
G, Teflon- lined
cap
G
P, G
G, Teflon- lined
cap
G, Teflon-lined
cap
G, Teflon- lined
cap
G, Teflon-lined
cap
G, Teflon-lined
cap
G, Teflon- lined
septum
G, Teflon- lined
septum
G, Teflon- lined
cap

P, G
None required
Cool, 4°C
Cool, 4°C
Coot, 4°C, add zinc acetate

Cool, 4°C
HNO, to pH<2
HN03 to pH<2


Cool, 4°C, 0.008% Na2S2033,
Adjust pH to 4-5
Cool, 4°C, 0.008% Na2S2033,
Adjust pH to 6-9, store in
dark
Cool, 4°C, 0.008% Na2S20,
n
Cool, 4°C, 0.008% Na2S203

Cool, 4°C, 0.008% Na2S2033

Cool, 4°C, 0.008% Na2S2033
store in dark
Cool, 4°C, 0.008% Na2S203,
store in dark
Cool, 4°C,
Cool, 4°CZ
Cool, 4°C

Cool, 4°C

Cool, 4°C, 0.008% Na2S2033

Cool, 4°C

Cool, 4°C, 0.008% Na2S2033
store in dark
Cool, 4°C, 0.008% Na2S203 '3

Cool, 4°C, 0.008% Na2S2033

Cool, 4°CZ


HN03 to pH<2
24 hours
48 hours
28 days
7 days

24 hours
28 days
6 months


14 days

7 days until extraction.
after extraction

7 days until extraction.
after extraction
7 days until extraction.
after extraction
7 days until extraction,
after extraction
7 days until extraction,
after extraction
7 days until extraction,
after extraction
28 days
28 days
7 days until extraction.
after extraction
7 days until extraction,
after extraction
7 days until extraction,
after extraction
7 days until extraction,
after extraction
7 days until extraction,
after extraction
14 days

14 days

28 days


6 months












40 days


40 days

40 days

40 days

40 days

40 days



40 days

40 days

40 days

40 days

40 days









1
Table excerpted,  in part,  from Table  II, 49  FR 209, October 26, 1984, p 28.
Polyethylene (P)  or Glass  (G)
Adjust to pH<2 with H2SO,,  HCl  or solid NaHSO,.
Free chlorine must be removed  prior to addition of HCl by the appropriate addition of  Na2S203.
                                                        TWO -  41
                                                                                                   Revision 2
                                                                                              September 1994

-------
                  TABLE 2-34.  PREPARATION METHODS FOR ORGANIC ANALYTES

Acids
Acrolein
Acrylonitrile
Acetonitrile
Aromatic Volatiles
Base/Neutral
Chlorinated
Herbicides
Chlorinated
Hydrocarbons
Halogenated
Volatiles
Nitroaromatic and
Cyclic Ketones
Non-halogenated
Volatiles
Organochlorine
Pesticides and PCBs
Organophosphorus
Pesticides
Phenols
Phthalate Esters
Polynuclear
Aromatic
Hydrocarbons
Volatile Orqanics
Aqueous (pH)3
3510
3520
(PH <2)
5030
5030
3510
3520
(pH >11)
8150
8151
(PH <2)
3510
3520
(PH 7)
5030
3510
3520
(pH 5-9)
5030
3510
3520
3665
(pH 5-9)
3510
3520
(pH 6-8)
3510
3520
(PH <2)
3510
3520
(pH 7)
3510
3520
(PH 7)
5030
Solids
3540, 3541
3550
35802
5030
5030
3540
3541
3550
35802
8150
8151,
35802
3540
3541
3550
35802
5030
3540
3541
3550
35802
5030
3540
3541
35802
3665
3540
3541
35802
3540
3541
3550
35802
3540
3541
3550
35802
3540, 3541
3550
35802
5030
Sludges
Emulsions1 (pH)
3520
(pH <2)
5030
5030
3520
(pH >11)
8150
8151
(PH <2)
3520
(pH 7)
5030
3520
(pH 5-9)
5030
3520
(pH 5-9)
3520
(pH 6-8)
3520
(pH <2)
3520
(pH 7)
3520
(pH 7)
5030
Oils
3650
35802
5030
5030
3650
35802
35802
35802
5030
35802
5030
35802
35802
3650
35802
35802
3560
35802
5030
1  If attempts  to break up emulsions are unsuccessful,  these  methods  may be used.

2  Method 3580  is only appropriate if the sample is  soluble  in  the  specified solvent.
3
  pH at which extraction should be performed.
                                         TWO - 42
    Revision 2

September 1994

-------
            TABLE 2-35.
CLEANUP OF ORGANIC ANALYTE EXTRACTS
Analyte Type
Acids
Base/Neutral
Chlorinated
Herbicides
Chlorinated
Hydrocarbons
Nitroaromatics &
Cyclic Ketones
Organophosphorus
Pesticides
Organochlorine
Pesticides &
PCBs
Phenols
Phthalate
Esters
Polynuclear
Aromatic
Hydrocarbons
Method(s)
3650
3650
8150
8151
3620
3640
3620
3640
3620
3620
3630
3640
3660
3665
3630
3640
3650
3610
3611
3620
3640
3610
3611
3630
3640
             TWO - 43
    Revision 2
September 1994

-------
                                      TABLE  2-36.
                         DETERMINATION OF  ORGANIC  ANALYTES

GC/MS Determination
Methods
Specific GC Detection
Methods
HPLC
SEMIVOLATILES
Acids
Base/Neutral
Carbamates
Chlorinated Herbicides
Chlorinated Hydrocarbons
Dyes
Explosives
Haloethers
Nitroaromatics and Cyclic
Ketones
Nitrosoamines
Organochlorine Pesticides and
PCBs
Organophosphorous Pesticides
Phenol s
Phthalate Esters
Polynuclear Aromatic
Hydrocarbons
8270
8250
8270
8250

8270*
8270
8250


8270
8250
8270
8250
8270
8250
8270*
8270*
8270
8250
8270
8250
8270
8250



8150
8151
8120
8121


8110
8090
8070
8080
8081
8140
8141
8040
8060
8061
8100


8318


8321
8330
8331




8321


8310
VOLATILES
Acrolein, Acrylonitrile,
Acetonitrile
Acryl amide
Aromatic Volatiles
Formaldehyde
Halogenated Volatiles
Non-halogenated Volatiles
Volatile Organics
8240
8260

8240
8260

8240
8260
8240
8240
8260
8030
8031
8032
8020
8021

8010
8011
8021
8015
8010
8011
8020
8021
8030
8031
8316
8315
8316

8315


8315
8316
*This method is an alternative confirmation method.  It is not the method of choice.
                                        TWO -  44
     Revision  2
September 1994

-------
  OO
  o
  I—I
  a.
 .O

-------
                                 FIGURE  2-2.
                  SCHEMATIC OF  SEQUENCE  TO  DETERMINE
              IF A WASTE  IS  HAZARDOUS  BY  CHARACTERISTIC
                                                DOT (49 CFR 173 300)
   Is
  waste
reactive to
air and/or
 water?
C                                                                         hazardous by\
                                                                         reason of     i
                                                                         ignitability    J
                                                                         laracteristic^/
      Is waste
     explosive?
                                                Generator Knowledge
                                                DOT (49 CFR 173.151)
       What is
     physical state
      of waste?
                                     Is waste
                                    ignitable?
Hazardous^)
     Perform Paint
      Filter Test
    (Method 9095)
             Methods 1110 and 9040

                    Yes
  Is waste
 corrosive?
/"   Nonhazardous  ^
(     for corrosivity    )
\c;haracteristi(;/
               Methods 1010 or 1020

                    Yes	
                                   TWO  - 46
                                                                          Revision  2
                                                                     September 1994

-------
        FIGURE  2-2.
        (Continued)
    Nonhazardous
    for ignitability
    characteristic
             Reactive CN
             and Sulfide Tests
     Does waste
    generate toxic
       gas?
   Nonhazardous
for toxic gas generation
(reactivity) characteristic
    Is total
   concen. of TC
   constituents-r 20 <
   TC regulatory
       limit?
     Is waste
   leachable and
      toxic?
   (Method 1311)
Nonhazardous
  for toxicity
characteristic
Nonhazardous^N.
  for toxicity      J
 characteristic^/
           TWO  -  47
                       Revision  2
                 September  1994

-------
                                FIGURE 2-3A.
                                     EP
                                  Sample
                                   1310
    3010
  (7760  Ag)
    6010
               7470
                Hg
   3510
 Neutral
   8150
   8151
Herbicides
 Ba--




Cr --


Ag --
-- As




-- Cd


-- Pb


-- Se
   8080
   8081
Pesticides
                                  TWO - 48
                                                            Revision 2
                                                        September 1994

-------
                       FIGURE 2-3B.
RECOMMENDED SW-846 METHODS OF ANALYSIS FOR TCLP LEACHATES





3010




6010



Ba -

Cr -

Ag -








- As

- Cd

- Pb

- Se


1

7470
Hg













Sample
. 	

TCLP












1

3510
Neutral






8240 3510
8260 (Acidic
Volatile and
Organics Basic)

, 	
8080
8081
Pestic-
ides








8270
Semi vol-
atile
Orqanics
















1
8150
8151
Herbic-
ides











                        TWO - 49
    Revision 2
September 1994

-------
                                 FIGURE 2-4A.
                            GROUND WATER ANALYSIS
Organic
Sample

1
VOA
, * ,
8240 or
8260


1
Semivolatiles
, I
3510 or
3520
, I
8270 or
8250



i

\ ' \ t
Pesticides
1
Herbicides Dioxins
1 1
3510 or
3520
Neutral
I

1 3620, 3640
and/or 3660
\

8080
81 50 8280

1 -Optional: Cleanup required only if interferences prevent analysis
                                    TWO - 50
     Revision 2
September  1994

-------
   FIGURE 2-4B.
INDICATOR  ANALYTE
 Indicator
 Analyte(s)
1
POC
                  1 - Barcelona, 1984, (See Reference 1)
                  2 - Riggin, 1984, (See Reference 2)
      TWO -  51
     Revision  2
September 1994

-------
                                   FIGURE  2-4C.
                                   GROUND  WATER
                              (   GROUND WATER1   )
                              V      SAMPLE      J
             SAMPLE PREPARATION
                 3005 OR 3015
SAMPLE PREPARATION
    3015 OR 3020
                      I
Ag, Al, As, Ba, Be,
Cd, Co, Cr, Cu, Fe,
Mg, Mn, Mo, Ni, Pb,
Sb, Se, Ti, V, Zn

Ag, Al, As, Ba, Be,
Cd, Co, Cr, Cu, Mn,
Ni, Pb. Sb, TI, Zn
Ag - 7760
Ba - 7080
Cd-7130
Cr-7190
Fe -7380
Mn - 7460
Ni - 7520
Sb - 7040
TI - 7840
Zn - 7950
Al - 7020
Be -7090
Co - 7200
Cu-7210
Mg - 7450
Mo - 7480
Pb - 7420
Sn - 7870
V - 7910
        i
Ag- 7761'
Ba - 7081 *
Be - 7091
Cd-7131
Co - 7201
Cr-7191
Cu - 721 1*
Fe-7381*
Mn-7461'
Mo - 7481
Pb - 7421
TI - 7841
Sb-7041*
7062*
V-7911
Zn - 7951*
*  Follow the digestion procedures as detailed in the individual
  determinative methods.

1  When analyzing for total dissolved metals, digestion is not
  necessary if the samples are filtered at the time of
  collection, and then acidified to the same concentration as the standards
                                       TWO  -  52
                            Revision  2
                      September  1994

-------
                                 CHAPTER THREE

                               METALLIC ANALYTES
3.1  SAMPLING CONSIDERATIONS

3.1.1  Introduction

      This manual contains procedures for the analysis of metals in a variety of
matrices.  These methods  are written  as  specific steps in the overall analysis
scheme -- sample handling  and preservation, sample digestion  or preparation, and
sample analysis for specific metal components.   From  these methods,  the analyst
must assemble a total  analytical protocol which is appropriate for the sample to
be analyzed and for the information required.  This  introduction discusses the
options  available  in  general  terms,  provides  background  information  on the
analytical techniques, and highlights some of the considerations to be made when
selecting a total analysis protocol.


3.1.2  Definition of Terms

      Optimum concentration  range:    A range,  defined by limits expressed in
concentration, below which  scale  expansion  must  be used  and above which curve
correction should be considered.  This range will vary with the  sensitivity of
the instrument and the operating conditions employed.

      Sensitivity:  a) Atomic  Absorption:   The  concentration in milligrams of
metal per liter that produces an absorption of 1%; b)  Inductively  Coupled Plasma
(ICP):   The  slope  of the analytical  curve,  i.e.,  the functional relationship
between emission intensity and concentration.

      Method detection limit  (MDL):   The minimum concentration  of a  substance
that  can be  measured  and  reported  with  99%  confidence   that the analyte
concentration is greater  than  zero.   The MDL  is  determined  from analysis of a
sample in a given matrix containing analyte which has  been processed  through the
preparative procedure.

      Total  recoverable metals:   The  concentration of metals in an unfiltered
sample following treatment with hot dilute mineral acid (Method 3005).

      Dissolved metals: The concentration of metals determined in a sample after
the sample is filtered through a 0.45-um filter (Method 3005).

      Suspended metals:  The concentration of metals determined  in the
portion of a sample that is retained by a 0.45-um filter (Method 3005).

      Total  metals:  The concentration of metals determined in a sample  following
digestion by Methods 3010, 3015, 3020, 3050 or 3051.
                                   THREE  -  1                       Revision 2
                                                                  September 1994

-------
      Instrument detection limit (IDL):  The concentration equivalent to a signal
due to the  analyte  which is equal to three times  the  standard deviation of a
series of  7 replicate measurements  of a reagent  blank's  signal  at  the same
wavelength.

      Interference check sample (ICS):  A solution containing both interfering
and analyte elements of known concentration that can be used  to verify background
and interelement correction factors.

      Initial  calibration  verification  standard  (ICV):     A  certified  or
independently  prepared  solution  used  to  verify the  accuracy of  the initial
calibration.   For ICP analysis, it  must be  run  at  each wavelength used in the
analysis.

      Continuing calibration  verification  (CCV):   Used  to assure calibration
accuracy during each analysis  run.   It must be run for each  analyte as described
in the particular analytical method.  At a minimum,  it  should be analyzed at the
beginning of the run and  after  the  last analytical sample.  Its concentration
should be at or near the mid-range levels of the calibration curve.

      Calibration standards:  A series of known standard solutions used by the
analyst for calibration of the instrument (i.e., preparation of the analytical
curve).

      Linear dynamic range:  The concentration range over which the analytical
curve remains linear.

      Method blank:  A  volume of reagent water processed  through each sample
preparation procedure.

      Calibration blank:   A volume  of  reagent  water acidified  with  the same
amounts of acids as were the standards and samples.

      Laboratory control  standard:  A volume of reagent water spiked with known
concentrations of  analytes and carried through  the preparation  and  analysis
procedure as a sample.   It is used to monitor loss/recovery values.

      Method  of  standard  addition  (MSA):    The   standard-addition  technique
involves the use of the  unknown and the unknown plus  several  known amounts of
standard.  See Method 7000, Section 8.7 for detailed instructions.

      Sample holding time:  The storage time allowed between sample collection
and sample analysis  when  the designated  preservation and storage techniques are
employed.
                                   THREE  -  2                       Revision 2
                                                                  September 1994

-------
      3.1.3  Sample Handling and Preservation

      Sample holding times, digestion procedures and suggested collection volumes
are listed  in  Table 1.   The sample volumes required depend upon the number of
different digestion procedures necessary for analysis.  This may be determined
by the  application  of graphite-furnace atomic  absorption spectrometry (GFAA),
flame atomic absorption  spectrometry (FLAA),  inductively coupled argon plasma
emission spectrometry  (ICP), hydride-generation atomic absorption spectrometry
(HGAA),  inductively  coupled plasma  mass  spectrometry  (ICP-MS)  or cold-vapor
atomic  absorption  spectrometry (CVAA)  techniques,  each of  which  may require
different digestion procedures.  The indicated volumes in Table 3-1 refer to that
required  for  the  individual   digestion   procedures and  recommended  sample
collection volumes.

      In the  determination of  trace metals,  containers can introduce either
positive  or  negative  errors   in  the  measurement  of  trace   metals  by  (a)
contributing  contaminants  through leaching or  surface  desorption,   and  (b)
depleting concentrations through adsorption.  Thus  the collection and treatment
of the  sample  prior to analysis require particular  attention.   The following
cleaning treatment  sequence has  been determined  to be adequate  to  minimize
contamination  in  the  sample  bottle,  whether  borosilicate  glass,  linear
polyethylene, polypropylene, or Teflon: detergent,  tap water, 1:1 nitric acid,
tap water,  1:1 hydrochloric acid,  tap water, and reagent water.

            NOTE: Chromic acid  should not be used to clean glassware, especially
            if chromium is to be included in the analytical  scheme.  Commercial,
            non-chromate  products  (e.g.,  Nochromix)  may  be used  in  place of
            chromic acid  if adequate cleaning  is  documented by an analytical
            quality control program.   (Chromic acid should also not  be used with
            plastic bottles.)

      3.1.4  Safety

      The toxicity or carcinogenicity of each reagent used  in these methods has
not been precisely defined.  However, each chemical compound should be treated
as a potential  health  hazard.   From this viewpoint, exposure  to these chemicals
must be reduced to the lowest possible level by whatever means available.  The
laboratory  is  responsible  for  maintaining a  current awareness file  of OSHA
regulations regarding  the safe handling of  the chemicals  specified  in  these
methods. A reference  file of material  data-handling  sheets should also be made
available  to  all  personnel  involved  in  the  chemical  analysis.   Additional
references  to laboratory safety are available.   They are:

1.    "Carcinogens  -  Working  with Carcinogens,"  Department of Health,
Education,  and Welfare,  Public Health Service, Center  for Disease
Control, National  Institute for Occupational  Safety and  Health,
Publication No. 77-206, August 1977.
                                   THREE  - 3                       Revision 2
                                                                  September 1994

-------
                                 TABLE 3-1.

  SAMPLE  HOLDING TIMES,  REQUIRED DIGESTION VOLUMES AND RECOMMENDED COLLECTION
         VOLUMES FOR METAL DETERMINATIONS IN AQUEOUS AND SOLID SAMPLES
Measurement
Digestion
Vol. Req.a
 (mL)
Collection
Volume (mL)a
Treatment/
Preservative
Holding Time0
Metals (except hexavalent chromium and mercury):
Aqueous
      Total
      Dissolved



      Suspended

Solid
      Total

Chromium VI:b

      Aqueous

      Solid

Mercury:

Aqueous
      Total


      Dissolved
    100
    100
    100
    2g
    100
Solid
      Total
    100
    100
    0.2g
  600
  600
  600
  200g
  400
  400
  200g
HN03 to pH <2
6 months

Filter on site;
HN03 to pH <2
6 months

Filter on site
6 months

6 months
  400          24 hr

  200g
HN03 to pH <2
28 days

Filter;
HN03 to pH <2
28 days

28 days
"Unless stated otherwise.
bThe holding time for the analysis of hexavalent chromium in solid samples has
not yet been determined.  A holding time of  "as soon as possible"  is recommended.
CA11  non-aqueous samples and  all  aqueous samples that  are  to  be analyzed for
mercury  and  hexavalent chromium must  be stored at  4°C  ±  2°C  until  analyzed,
either glass or plastic containers may  be used.
                                   THREE - 4
                                       Revision  2
                                       September 1994

-------
2.        "OSHA Safety and Health Standards, General Industry" (29 CFR  1910),
Occupational  Safety  and Health Administration, OSHA  2206,  revised
January 1976.

3.        "Proposed OSHA Safety and Health Standards, Laboratories," Occupational
Safety and Health Administration, Federal Register, July 24,  1986, p. 26660.

4.        "Safety in  Academic Chemistry Laboratories,"  American Chemical  Society
Publication,  Committee  on Chemical Safety, 3rd edition, 1979.
3.2  SAMPLE PREPARATION METHODS

          The methods  in  SW-846  for  sample digestion  or preparation  are as
follows1:

          Method 3005 prepares ground  water  and  surface water  samples for total
recoverable and dissolved  metals  determination by FLAA, ICP-AES, or ICP-MS.  The
unfiltered or filtered sample is heated with dilute HC1  and HN03 prior to metal
determination.

          Method 3010 prepares waste  samples for  total  metal  determination by
FLAA, ICP-AES, or ICP-MS.   The samples are vigorously  digested with nitric  acid
followed by dilution with hydrochloric acid.  The method is applicable  to aqueous
samples, EP and mobility-procedure extracts.

          Method 3015 prepares aqueous samples,  mobility-procedure  extracts, and
wastes that contain  suspended solids for total metal determination by FLAA, GFAA,
ICP-AES, or  ICP-MS.  Nitric  acid is  added to  the  sample in a Teflon digestion
vessel and heated in a microwave unit prior to metals determination.

          Method 3020 prepares waste samples for total metals determination by
furnace GFAA or  ICP-MS.   The  samples  are vigorously digested with nitric  acid
followed by  dilution with nitric  acid.   The method  is  applicable to aqueous
samples, EP and mobility-procedure extracts.

          Method 3040 prepares oily waste samples  for determination of soluble
metals by FLAA,  GFAA, and ICP-AES methods. The samples are dissolved and diluted
in organic solvent prior to analysis.  The method is applicable to the organic
extract in the oily  waste  EP procedure and other samples  high  in oil, grease, or
wax content.

          Method 3050 prepares waste samples for total metals determination by
FLAA and ICP-AES, or ICP-MS.  The samples are vigorously  digested in nitric  acid
and hydrogen peroxide followed by  dilution  with either  nitric or hydrochloric
acid.  The method is applicable to soils, sludges, and solid waste samples.

          Method 3051 prepares  sludges,   sediments,  soils  and oils  for  total
metals determination by FLAA,  GFAA, ICP-AES  or  ICP-MS.  Nitric acid is added to


                                   THREE  - 5                      Revision 2
                                                                  September 1994

-------
the representative sample in a Teflon digestion vessel and heated in a microwave
unit prior to metals determination.

1  Please note that  chlorine is  an interferent in  ICP-MS analyses and  its  use
should be discouraged except when absolutely necessary.
                                   THREE - 6                       Revision  2
                                                                   September 1994

-------
                                 METHOD 3005A

               ACID  DIGESTION OF WATERS  FOR  TOTAL  RECOVERABLE  OR
           DISSOLVED METALS FOR ANALYSIS BY  FLAA OR  ICP  SPECTROSCOPY
1.0  SCOPE AND APPLICATION

      1.1   Method 3005 is an acid digestion procedure used to prepare surface
and ground water  samples  for  analysis  by flame atomic absorption spectroscopy
(FLAA) or  by inductively  coupled argon  plasma  spectroscopy (ICP).   Samples
prepared by Method 3005 may be analyzed by AAS  or ICP for the following metals:

            Aluminum                      Magnesium
            Antimony**                    Manganese
            Arsenic*                      Molybdenum
            Barium                        Nickel
            Beryllium                     Potassium
            Cadmium                       Selenium*
            Calcium                       Silver
            Chromium                      Sodium
            Cobalt                        Thallium
            Copper                        Vanadium
            Iron                          Zinc
            Lead

            * ICP only
            **May be analyzed by  ICP, FLAA,  or GFAA

      1.2   When analyzing for total  dissolved  metals  filter the sample, at the
time of collection, prior to acidification with nitric acid.


2.0  SUMMARY OF METHOD

      2.1   Total recoverable metals  - The entire sample is acidified at the time
of collection with nitric  acid.   At  the time of analysis  the sample is heated
with acid and substantially reduced  in  volume.   The digestate is filtered and
diluted to volume, and is then ready for analysis.

      2.2   Dissolved metals - The sample is filtered through a 0.45-/zm filter
at the time of collection and  the liquid phase  is then acidified at the time of
collection with  nitric  acid.   Samples  for dissolved metals do  not  need  to be
digested as  long as the  acid concentrations  have  been  adjusted to  the  same
concentration as in the standards.
3.0  INTERFERENCES

      3.1   The analyst should be cautioned that this digestion procedure may not
be sufficiently vigorous to destroy some metal complexes.
                                   3005A  -  1                       Revision 1
                                                                  July 1992

-------
Precipitation will  cause a lowering of the silver concentration and therefore an
inaccurate analysis.


4.0  APPARATUS AND MATERIALS

      4.1   Griffin beakers of assorted sizes or equivalent.

      4.2   Watch glasses or equivalent.

      4.3   Qualitative filter paper and filter funnels.

      4.4   Graduated cylinder or equivalent.

      4.5   Electric  hot  plate  or  equivalent  -  adjustable  and capable  of
maintaining a temperature of 90-95°C.


5.0  REAGENTS

      5.1   Reagent grade chemicals  shall be used in all  tests. Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications
of the Committee on Analytical  Reagents of the American  Chemical Society, where
such specifications are  available. Other grades  may  be used, provided it  is first
ascertained that the reagent  is  of  sufficiently high  purity to  permit  its use
without lessening the accuracy of the determination.

      5.2   Reagent Water.   Reagent  water  shall  be  interference free.   All
references  to water in the method  refer to  reagent water  unless  otherwise
specified.  Refer to Chapter One for a definition of reagent water.

      5.3   Nitric  acid  (concentrated),  HNO,.   Acid  should  be  analyzed  to
determine level of impurities.  If method blame is < MDL, then  acid  can  be used.

      5.4   Hydrochloric acid  (concentrated), HC1.   Acid should be analyzed to
determine level of impurities.  If method blank is < MDL, then  acid  can  be used.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples must have  been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with detergents, acids, and
water.  Both  plastic and glass containers are suitable.

      6.3   Sampling

            6.3.1   Total  recoverable metals - All  samples must  be  acidified at
      the time of collection with HN03 (5 ml/I).

            6.3.2   Dissolved metals  - All  samples must be filtered through a
      0.45-/im filter and  then  acidified  at the  time of  collection  with HNO,
      (5 ml/I).

                                   3005A -  2                        Revision 1
                                                                  July 1992

-------
7.0  PROCEDURE

      7.1   Transfer a 100-mL aliquot of well-mixed sample to a beaker.

      7.2   For metals that are to be analyzed,  add 2 mL of concentrated HN03 and
5 ml of concentrated HC1.   The  sample  is  covered with  a ribbed watch glass or
other suitable  covers  and  heated on a steam bath, hot  plate  or other heating
source at 90 to 95°C until  the volume has  been  reduced  to 15-20 ml.

      CAUTION:    Do not boil.  Antimony is easily lost by volatilization from
                  hydrochloric acid media.

      7.3   Remove the beaker and allow to cool.  Wash down the  beaker walls and
watch glass with water and, when necessary, filter or centrifuge the sample to
remove silicates  and  other insoluble material  that  could clog the nebulizer.
Filtration should be done only if there  is concern that  insoluble materials may
clog the nebulizer;  this  additional step is liable to  cause sample contamination
unless the filter and filtering apparatus are thoroughly cleaned and prerinsed
with dilute HN03.

      7.4   Adjust the final volume to 100 ml with reagent water.


8.0  QUALITY CONTROL

      8.1   All quality  control  measures described  in Chapter One  should be
fol1 owed.

      8.2   For each  analytical  batch  of samples processed,  blanks  should be
carried throughout the entire sample preparation and  analytical  process.  These
blanks will be useful in determining if samples are being contaminated.  Refer
to Chapter One for the proper protocol  when analyzing blanks.

      8.3   Replicate  samples should  be  processed  on  a routine  basis.    A
replicate sample  is a  sample  brought through the  whole sample preparation and
analytical process.   Replicate samples will be used to determine precision.  The
sample load will dictate  the frequency, but 5% is  recommended.   Refer to Chapter
One for the proper protocol when analyzing replicates.

      8.4   Spiked samples or  standard reference materials should be employed to
determine accuracy.   A spiked  sample should be  included  with each batch.  Refer
to Chapter One for the proper protocol  when analyzing spikes.


9.0  METHOD PERFORMANCE

      9.1   No data provided.
                                   3005A  -  3                       Revision 1
                                                                  July 1992

-------
10.0  REFERENCES

1.    Rohrbough,  W.G.;  et  al.  Reagent  Chemicals.  American  Chemical  Society
Specifications, 7th ed.; American Chemical Society: Washington, DC, 1986.

2.    1985 Annual  Book of ASTM Standards.  Vol. 11.01; "Standard  Specification for
Reagent Water"; ASTM: Philadelphia, PA,  1985; D1193-77.
                                                                                   i
                                   3005A - 4
Revision 1
July 1992

-------
                          METHOD 3005A
     ACID DIGESTION OF  WATERS  FOR TOTAL RECOVERABLE  OR
DISSOLVED METALS FOR ANALYSIS BY  FLAA OR  ICP SPECTROSCOPY
                           Start
                        7.1 Tr«n»fmr
                         aliquot of
                          (ample to
                          beaker
                           7 2 Add
                        concentrated
                        HNO, and HC1
                          7.2 H.at
                          •ample to
                        reduce volume
                          7.3 Cool
                          beaker;
                          filter if
                          necenary
                         7.4 Adjutt
                        final volume
                           Stop
                          3005A  -  5
Revision 1
July  1992

-------
                                 METHOD 3010A

              ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR
             TOTAL  METALS  FOR ANALYSIS  BY  FLAA  OR  ICP  SPECTROSCOPY
1.0  SCOPE AND APPLICATION

      1.1   This digestion  procedure is used  for the preparation  of aqueous
samples, EP and mobility-procedure extracts, and wastes that contain suspended
solids  for  analysis,   by  flame  atomic  absorption  spectroscopy  (FLAA)  or
inductively coupled argon plasma spectroscopy (ICP).  The procedure is used to
determine total metals.

      1.2   Samples prepared by Method 3010 may be analyzed by FLAA or ICP for
the following:

            Aluminum                      Magnesium
            *Arsenic                      Manganese
            Barium                        Molybdenum
            Beryllium                     Nickel
            Cadmium                       Potassium
            Calcium                       *Selenium
            Chromium                      Sodium
            Cobalt                        Thallium
            Copper                        Vanadium
            Iron                          Zinc
            Lead

      * Analysis by ICP

      NOTE: See Method 7760 for the digestion and FLAA analysis of Silver.

      1.3   This digestion procedure is not suitable for samples which will be
analyzed by graphite furnace atomic absorption spectroscopy because hydrochloric
acid can cause interferences during furnace atomization.   Consult Method 3020A
for samples requiring graphite furnace analysis.


2.0   SUMMARY OF METHOD

      2.1   A mixture of nitric acid and the material to be analyzed  is refluxed
in a covered Griffin  beaker.   This  step  is repeated with additional portions of
nitric  acid until  the  digestate  is  light  in  color or  until  its  color has
stabilized.  After  the digestate has been brought to a low  volume,  it  is refluxed
with hydrochloric acid  and brought up to  volume.  If sample should  go  to dryness,
it must be discarded and the sample reprepared.


3.0   INTERFERENCES

      3.1   Interferences are discussed in the referring  analytical  method.
                                   3010A -  1                       Revision 1
                                                                  July 1992

-------
4.0   APPARATUS AND MATERIALS

      4.1   Griffin beakers - 150-mL or equivalent.

      4.2   Watch glasses - Ribbed and plain or equivalent.

      4.3   Qualitative filter paper or centrifugation equipment.

      4.4   Graduated cylinder or equivalent - lOOmL.

      4.5   Funnel or equivalent.

      4.6   Hot plate or equivalent heating source - adjustable and capable of
maintaining a temperature of 90-95°C.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform to the specifications
of the Committee on Analytical Reagents of the American Chemical Society, where
such specifications  are  available.   Other grades may be  used,  provided it is
first ascertained that the reagent is  of sufficiently high  purity to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent  Water.    Reagent  water  will  be  interference  free.   All
references  to  water  in  the  method  refer to  reagent water  unless  otherwise
specified.  Refer to Chapter One for a definition of reagent water.

      5.3   Nitric  acid  (concentrated),  HN03.    Acid should  be  analyzed  to
determine levels of impurities.   If method blank is < MDL,  the  acid  can be used.

      5.4   Hydrochloric acid (1:1), HC1.  Prepared from water and  hydrochloric
acid.  Hydrochloric  acid  should  be analyzed to determine  level  of impurities.
If method blank is < MDL, the acid can be used.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must have  been  collected using  a  sampling  plan that
addresses the considerations  discussed  in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with detergents, acids, and
water.  Plastic and glass  containers are both suitable. See Chapter Three, Step
3.1.3, for further information.

      6.3   Aqueous  wastewaters must be  acidified to  a pH  of < 2 with HN03.


7.0   PROCEDURE

      7.1   Transfer a 100-mL representative aliquot of the well-mixed  sample to
a 150-mL Griffin beaker and  add 3 mL of concentrated HNO,.   Cover the beaker with
a  ribbed watch  glass or  equivalent.    Place the  bealcer  on  a hot  plate or

                                   3010A -  2                       Revision 1
                                                                   July  1992

-------
equivalent heating source and cautiously evaporate to a low volume (5 ml), making
certain that the sample does not boil and that no portion of the bottom of the
beaker is allowed to go dry.   Cool  the  beaker and  add  another 3-mL  portion of
concentrated HNQj.   Cover the  beaker with a nonribbed watch glass and return to
the hot plate,   increase the temperature of the hot plate so that  a gentle reflux
action occurs.

      NOTE: If a sample is  allowed to go to dryness, low recoveries will result.
            Should this occur, discard the  sample and reprepare.

      7.2   Continue heating,  adding additional  acid  as  necessary,  until  the
digestion is complete (generally indicated when the digestate  is light in color
or does not change in appearance with continued refluxing).  Again, uncover the
beaker or use a ribbed watch  glass,  and evaporate  to  a low volume (3 ml),  not
allowing any portion of the bottom  of the beaker to  go dry.   Cool the beaker.
Add a  small  quantity of 1:1  HC1  (10 mL/100  ml of final  solution),  cover  the
beaker, and reflux for an additional 15 minutes to dissolve any precipitate or
residue resulting from evaporation.

      7.3   Wash down  the  beaker walls  and  watch  glass with water  and,  when
necessary,  filter or  centrifuge  the sample  to remove  silicates   and  other
insoluble material  that could clog the nebulizer.  Filtration should be done only
if  there  is concern that  insoluble materials may clog  the  nebulizer.   This
additional step can cause sample contamination unless the filter and filtering
apparatus are thoroughly cleaned.   Rinse the filter  and  filter apparatus with
dilute nitric acid and discard  the  rinsate.   Filter  the  sample and  adjust  the
final  volume to 100 ml with reagent water and the  final  acid concentration to
10%.  The sample is now ready for analysis.


8.0   QUALITY CONTROL

      8.1   All quality  control measures described  in Chapter One  should be
followed.

      8.2   For each  analytical  batch of samples processed,  blanks  should be
carried throughout the entire  sample-preparation  and analytical  process.  These
blanks will be useful in determining if samples are being contaminated.  Refer
to Chapter One for the proper protocol when  analyzing blanks.

      8.3   Replicate  samples  should  be processed  on  a routine  basis.    A
replicate sample is  a  sample  brought through the whole sample preparation  and
analytical process.  A  replicate  sample should be  processed with  each analytical
batch or every 20 samples,  whichever is greater.   Refer to Chapter One for the
proper protocol when analyzing replicates.

      8.4   Spiked samples  or  standard reference materials should be employed to
determine  accuracy.   A spiked  sample should be  included with each  batch of
samples processed and whenever a new sample matrix  is being analyzed.  Refer to
Chapter One for the proper protocol when analyzing spikes.

      8.5   The method of standard addition shall  be used for the analysis of all
EP extracts and delisting  petitions (see Method  7000,  Step 8.7).  Although  not
required, use of the method of standard addition is recommended for any sample

                                   3010A - 3                       Revision  1
                                                                  July 1992

-------
that is suspected of having an interference.


9.0   METHOD PERFORMANCE

      9.1   No data provided.


10.0  REFERENCES

1.     Rohrbough, W.G.;  et  al.  Reagent Chemicals,  American Chemical  Society
Specifications, 7th ed.; American Chemical Society: Washington,  DC, 1986.

2.     1985 Annual Book of ASTM Standards, Vol.  11.01; "Standard Specification for
Reagent Water"; ASTM: Philadelphia, PA, 1985; D1193-77.
                                   3010A -  4                      Revision 1
                                                                  July 1992

-------
                         METHOD 3010A
    ACID  DIGESTION  OF AQUEOUS  SAMPLES AND EXTRACTS
FOR TOTAL METALS ANALYSIS BY  FLAA OR ICP SPECTROSCOPY
                             Start
                       7.1 Trantfar *aapli
                       aliquot  to baakar,
                       add concantratad
                            HNO.
                          7.1 H.at to
                       avaporata to low
                       voluaa, cool, and
                       add concantratad
                            HNO,
                          7.1 Rahaat,
                           incraaaa
                        taaparatura to
                         craata gantla
                         raflux action
                          7.2 Haat to
                       complata digaation,
                         avaporata,add
                        KCl,warn baakar
                         7.3 Filtar if
                         nacaoary and
                         adjuat voluma
                           3010A  -  5
Revision  1
July 1992

-------
                                  METHOD 3015

                 MICROWAVE ASSISTED ACID DIGESTION OF AQUEOUS
                             SAMPLES AND EXTRACTS
 1.0  SCOPE AND APPLICATION

         1.1   This digestion procedure  is  used for the preparation of  aqueous
 samples, mobility-procedure extracts, and wastes that  contain  suspended solids
 for  analysis, by  flame  atomic  absorption spectroscopy  (FLAA),  graphite  furnace
 absorption spectroscopy (GFAA),  inductively coupled argon plasma spectroscopy
 (ICP),  or  inductively  coupled  argon  plasma mass  spectrometry (ICP-MS).  The
 procedure is a hot acid  leach for determining available  metals.  Due to the rapid
 advances  in  microwave   technology,  consult  your  manufacturer's  recommended
 instructions  for  guidance on their microwave digestion system  and refer to the
 SW-846  "DISCLAIMER" when conducting analyses using  Method 3015.

         1.2   Samples  prepared  by Method 3015 using nitric acid  digestion may be
 analyzed by FLAA,  GFAA,  ICP-AES, or ICP-MS  for the  following:

                          Aluminum             Lead
                          Antimony             Magnesium
                          Arsenic*             Manganese
                          Barium               Molybdenum
                          Beryllium            Nickel
                          Cadmium              Potassium
                          Calcium              Selenium*
                          Chromium             Silver
                          Cobalt               Sodium
                          Copper               Thallium
                          Iron                 Vanadium
                                               Zinc


                          *Cannot be analyzed by FLAA

 2.0  SUMMARY OF METHOD

         2.1   A  representative 45 mL  aqueous  sample  is  digested  in  5 mL of
 concentrated nitric acid in a  fluorocarbon  (PFA or  TFM) digestion vessel  for 20
minutes using microwave heating.   After the digestion process,  the sample is
 cooled, and then  filtered, centrifuged, or  allowed to  settle in a clean  sample
 bottle prior to analysis.

3.0  INTERFERENCES

         3.1   Many samples  that contain organics,  such as TCLP extracts, will
result in higher vessel  pressures which have the potential to cause venting of
the vessels.   Venting  can result in either loss  of analytes and/or sample, which
must be avoided.  A smaller sample size can be used but the final water volume

                                   3015 - 1                       Revision 0
                                                                  September 1994

-------
prior to nitric acid addition must remain  at  45 ml.  This is required to retain
the heat characteristics of the calibration procedure.  Limits of quantitation
will change with sample quantity  (dilution) as with instrumentation."

4.0  APPARATUS AND MATERIALS

        4.1  Microwave  apparatus  requirements

               4.1.1   The  microwave  unit  provides  programmable power  with  a
        minimum  of 574 W,  which  can  be  programmed to within  + 10 W  of the
        required  power.  Typical  units provide  a  nominal  600 W to  1200  W of
        power.   Temperature monitoring and control of  the  microwave unit are
        desirable.

               4.1.2   The  microwave  unit  cavity  is  corrosion  resistant and
        well ventilated.

               4.1.3   All  electronics  are  protected against  corrosion for safe
        operation.

               4.1.4   The  system requires  fluorocarbon  (PFA or  TFM)  digestion
        vessels  (120  mL capacity) capable of withstanding  pressures up to 7.5
        ± 0.7  atm  (110  ± 10 psig) and capable of controlled pressure relief at
        pressures  exceeding  7.5 ± 0.7  atm  (110 ± 10 psig).

               4.1.5   A rotating  turntable is employed to  insure  homogeneous
        distribution  of microwave radiation within the unit.  The speed of the
        turntable  should be  a minimum  of  3 rpm.

               CAUTION:  Those laboratories now using or contemplating  the use of
               kitchen type microwave  ovens for  this method  should  be aware of
               several significant safety  issues. First, when an acid  such as
               nitric  is used to assist sample digestion in  microwave units in
               open  vessels,  or  sealed  vessels equipped  with venting features,
               there is  the potential  for the  acid gases  released to corrode the
               safety  devices that prevent  the microwave magnetron from shutting
               off when  the door is opened.  This can  result in operator exposure
               to microwave energy. Use of a unit  with corrosion resistant safety
               devices prevents  this from  occurring.

               CAUTION:  The second safety  concern relates  to the use of sealed
               containers without  pressure  relief valves in the unit.  Tempera-
               ture is the  important variable controlling the  reaction.  Pressure
               is needed to attain  elevated temperatures  but must be safely con-
               tained.   However, many digestion vessels constructed from certain
               fluorocarbons  may  crack, burst,  or  explode  in the  oven under
               certain pressures.     Only  unlined  fluorocarbon  (PFA  or  TFM)
               containers  with  pressure relief  mechanisms  or containers  with
               fluorocarbon (PFA or TFM)  liners  and pressure relief mechanisms
               are  considered  acceptable at present.


                                   3015 -  2                       Revision 0
                                                                  September 1994

-------
               Users are therefore  advised  not to use  kitchen type microwave
               ovens or to  use  sealed containers without pressure relief valves
               for microwave acid digestions  by this method.  Use of laboratory
               grade microwave  equipment  is required to prevent safety hazards.
               For further  information  consult  reference  1.

               CAUTION: In  addition,   there  are many  safety  and  operational
               recommendations  specific  to the model  and manufacturer  of the
               microwave equipment  used  in  individual  laboratories.    These
               specific suggestions are  beyond the scope  of  this  method and
               require  the  analyst  to  consult the specific  equipment  manual,
               manufacturer and literature for  proper and safe  operation of the
               microwave equipment and  vessels.

         4.2 Volumetric graduated cylinder, 50  or 100 ml capacity or equivalent.

         4.3  Filter paper,  qualitative or equivalent.

         4.4  Analytical balance, 300 g capacity, minimum accuracy ±  0.01  g.

         4.5  Filter funnel, glass or disposable polypropylene.

5.0  REAGENTS

         5.1   Reagent  grade chemicals shall  be  used  in  all  tests.   Unless
otherwise  indicated,  it is  intended  that all  reagents  shall  conform  to the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where such  specifications  are available.   Other  grades  may be used,
provided it is first ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the  accuracy of the determination.   If the
purity of a reagent is  questionable, analyze  the reagent to determine the level
of impurities.  The reagent blank must  be less than the MDL  in order to be used.

         5.2  Reagent Water.   Reagent  water  shall  be interference  free.   All
references to water in the method refer  to reagent water unless otherwise specif-
ied (Ref. 2).

         5.3   Concentrated nitric  acid, HNO,.   Acid  should  be analyzed  to
determine levels of impurities.   If the method blank is less than the MDL, the
acid can be used.

6.0  SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

         6.1  All samples  must have been  collected using a  sampling plan that
addresses the considerations discussed in Chapter Nine  of this manual.

         6.2 All sample containers must be prewashed with detergents, acids, and
water.  Plastic containers are preferable. See Chapter Three,  Step 3.1.3 of this
manual, for further information.

         6.3 Aqueous waste  waters must be acidified to a pH of < 2 with HN03.

                                   3015 - 3                       Revision 0
                                                                  September 1994

-------
7.0  PROCEDURE

        7.1  Calibration of Microwave Equipment

              NOTE:  If the microwave unit  uses temperature feedback  control
              capable  of replicating  the performance  specifications  of  the
              method,  then the calibration procedure may be omitted.

              7.1.1  Measurement  of  the  available  power  for   heating   is
        evaluated so that absolute power  in watts may be  transferred from  one
        microwave unit to another.   For cavity  type microwave  equipment,  this
        is  accomplished  by measuring the  temperature  rise in  1  kg of  water
        exposed to microwave radiation for a fixed period of time.   The  analyst
        can relate power in watts to the partial  power setting of the unit.  The
        calibration format required  for laboratory microwave units  depends on
        the  type  of  electronic  system used  by the manufacturer to  provide
        partial microwave power.   Few units have an accurate and precise linear
        relationship between percent power settings and  absorbed power.  Where
        linear circuits have been  utilized, the calibration curve can be deter-
        mined  by  a  three-point  calibration  method (7.1.3),  otherwise,  the
        analyst must use the multiple point calibration method (7.1.2).

              7.1.2  The multiple  point  calibration involves the  measurement
        of absorbed power over a large range of power settings.   Typically,  for
        a 600 W unit,  the following power settings  are measured; 100,99,98,97,
        95,90,80,70,60,50,  and  40% using  the  procedure described  in  section
        7.1.4.  This data is clustered about the customary working power ranges.
        Nonlinearity  has  been commonly encountered  at  the upper end of  the
        calibration.   If the unit's  electronics are  known to have  nonlinear
        deviations  in  any region  of proportional  power  control,  it  will  be
        necessary to make  a  set  of measurements that bracket the power  to be
        used.   The final  calibration point  should  be at  the partial  power
        setting that will be used in the test.  This setting should  be  checked
        periodically  to evaluate  the integrity  of the  calibration.    If  a
        significant change  is detected (±10  W),  then  the entire  calibration
        should be reevaluated.

              7.1.3  The three-point  calibration  involves the measurement of
        absorbed power at three different  power settings.  Measure the power at
        100%  and  50%  using  the  procedure described  in  section  7.1.4,  and
        calculate the power  setting corresponding to the  required power  in watts
        specified  in   the  procedure  from  the  (2-point)  line.   Measure  the
        absorbed power at that partial power setting.  If the measured absorbed
        power does not correspond to the specified  power within ±10 W,  use  the
        multiple point calibration in 7.1.2.  This point should also be used to
        periodically verify the integrity of the calibration.

              7.1.4  Equilibrate a  large  volume of water to room temperature
        (23 ±2 °C).   One  kg of reagent water is weighed  (1,000.0  g ± 0.1 g)
        into a fluorocarbon (PFA or TFM)  beaker or a beaker made of some other
        material  that  does not significantly  absorb  microwave energy (glass

                                   3015  -  4                       Revision  0
                                                                  September  1994

-------
absorbs  microwave  energy  and  is  not   recommended).    The  initial
temperature of the water should be 23 ± 2 *C measured to ± 0.05  °C.  The
covered  beaker  is  circulated  continuously (in the normal sample path)
through  the microwave field for 2 minutes at the desired partial power
setting  with  the unit's exhaust fan on maximum  (as  it will  be during
normal  operation).   The  beaker is  removed  and the  water  vigorously
stirred.    Use   a  magnetic stirring  bar  inserted  immediately  after
microwave  irradiation  and record  the  maximum temperature within  the
first  30 seconds to ±  0.05 "C.  Use a new  sample  for each  additional
measurement.  If the water is  reused both  the  water and the beaker must
have returned to 23 ±  2 *C.   Three measurements at each power setting
should be made.

The absorbed power  is determined  by the following  relationship

                     P - (K)  (Cp)  (m)  (AT)

Eq. 1
Where:

P  =  the  apparent  power  absorbed  by  the  sample  in  watts  (W).
(W=joule-sec  )

K  =  the conversion factor  for  thermochemical  calories-sec"1  to watts
(=4.184)

C  * the heat capacity, thermal  capacity, or specific heat
(cal-g~1"C~1), of water

m  = the mass  of the water sample in grams (g)

AT = the final temperature minus the initial temperature  (°C)

t = the time  in seconds (s)

Using the  experimental  conditions  of 2 minutes and  1  kg  of distilled
water (heat  capacity  at 25 °C  is  0.9997  cal-g  •"C"1)  the calibration
equation simplifies to:

                       P = (AT)  (34.86)

      NOTE: Stable line voltage  is  necessary for accurate and reproduc-
      ible calibration and operation.  The line voltage  should be within
      manufacturer's specification, and during measurement and operation
      not  vary  by more than  ±2 V.   A constant  power supply  may  be
      necessary for microwave use if the source of the line voltage is
      unstable.
                           3015 -  5                       Revision 0
                                                          September 1994

-------
         Electronic components in most microwave units are matched to the units'
         function  and  output.  When any part of the high voltage circuit, power
         source,   or  control  components  in  the  unit  have  been   serviced  or
         replaced, it will be necessary to recheck the units' calibration power.
         If  the power  output has changed  significantly (±10 W), then the entire
         calibration  should be reevaluated.

         7.2  All  digestion vessels  and volumetric ware must be carefully acid
washed  and  rinsed with  reagent water.   When  switching between  high solids
(concentrated) samples and low solids (low concentration) samples all digestion
vessels should be cleaned by  leaching with hot  (1:1)  hydrochloric acid  (greater
than 80°C,  but less  than boiling)  for a minimum of two hours followed with hot
(1:1) nitric acid (greater than 80°C,  but  less than boiling) for a minimum of two
hours,  rinsed with  reagent  water,   and  dried  in a  clean environment.   This
cleaning procedure should also be used whenever the prior use of the digestion
vessels is unknown or cross contamination from  vessels is suspected.  Polymeric
or glass volumetric ware  and storage containers should  be cleaned by leaching
with more dilute acids  (approximately  10%  V/V)  appropriate  for  the specific
plastics used and then rinsed with reagent water and dried in a clean environ-
ment.   In addition, to  avoid precipitation  of  silver,  ensure that all  HC1 has
been rinsed from the vessels.

         7.3  Sample Digestion

              7.3.1   Weigh the fluorocarbon  (PFA or TFM)  digestion vessel, valve
         and cap  assembly to  0.01 g prior to  use.

              7.3.2   A  45 ml aliquot of  a well shaken sample is measured in a
         graduated cylinder.  This  aliquot is poured into the digestion vessel
         with  the  number of the  vessel  recorded on the preparation  sheet.

              7.3.3   A  blank sample  of  reagent water is treated  in  the same
         manner along  with spikes and duplicates.

              7.3.4   Add 5 mL of  concentrated  nitric acid to each vessel that
         will  be  used.   Check to make sure the  pressure relief disks are in the
         caps  with the smooth side toward the  sample and start the caps a few
         turns on  the vessels. Finish tightening the  caps  in the capping station
         which will tighten them to a uniform torque  pressure  of 12 ft-lbs.
         (16 N-m) or to the manufacturers  recommended specifications. Weigh each
         capped vessel to the nearest 0.01 g.

              CAUTION:  Toxic nitrogen oxide fumes may be  evolved, therefore all
              work must be performed  in a properly operating ventilation system.
              The analyst should also be aware  of the potential for a vigorous
              reaction.   If  a vigorous  reaction  occurs,  allow to cool before
              capping the vessel.

              7.3.5   Evenly  distributed  the vessels in the carousel according
         to  the manufacturer's recommended specifications.   Blanks are treated
         as  samples  for  the purpose  of balancing  the power  input.   When fewer

                                   3015  - 6                        Revision 0
                                                                   September 1994

-------
than  the recommended  number  of samples  are  digested,  the  remaining
vessels should be filled with  45 ml  of reagent water and 5 ml of nitric
acid to achieve the full compliment of vessels.  This provides an energy
balance since the microwave power absorbed is proportional to the total
mass  in the cavity  (Ref. 1).

      7.3.6  Program the microwave unit according to the manufacturer's
recommended specifications  and,  if  used,  connect  the  pressure vessels
to the central overflow vessel with  PFA-fluorocarbon tubes.  The chosen
sequence will  bring  the samples  to  160°C  ±  4°C  in 10  minutes and will
permit a slow rise to 165-170  °C during the second 10 minutes (Ref. 3).
Start the turntable  motor  and be sure the vent  fan  is running on high
and the turntable  is turning.   Start the microwave generator.

               7.3.6.1   Newer microwave  units  are  capable  of  higher
      power that  permit digestion  of a  larger  number of  samples per
      batch.   If the analyst wishes to digest more  samples  at a time,
      the analyst  may use  different  power settings  as  long as  they
      result  in  the same  time  and  temperature  conditions defined  in
      7.3.6.  That is,  any  sequence  of power that brings the samples to
      1608C ±  4°C  in 10 minutes and permits a slow rise  to  165-170eC
      during the second 10 minutes  (Ref.  2).

      Issues  of safety,  structural  integrity  (both  temperature  and
      pressure  limitations),   heat  loss,   chemical   compatibility,
      microwave absorption  of vessel material, and  energy transport will
      be considerations made in  choosing  alternative  vessels.   If all
      of the considerations are  met and the  appropriate power settings
      are provided to  reproduce the reaction  conditions defined  in
      7.3.6, then these alternative vessels may be used (Ref. 1,3)

      7.3.7   At  the end of the microwave program,  allow the  vessels
to cool  for at least  5  minutes in the  unit  before removal  to  avoid
possible injury if a vessel vents immediately  after  microwave heating.
The samples  may be  cooled outside the unit by removing the carousel and
allowing the samples to cool on  the  bench or in a water bath.   When the
vessels have cooled to  room temperature, weigh and record the weight of
each  vessel  assembly.    If the  weight  of  the  sample  plus  acid  has
decreased by more than 10% discard the sample.

      7.3.8  Complete  the   preparation  of  the  sample  by  carefully
uncapping and venting each  vessel in a fume  hood.  Transfer the  sample
to an acid-cleaned   bottle.    If  the digested  sample contains  par-
ticulates which may clog nebulizers  or interfere with  injection  of the
sample into the instrument, the  sample may  be centrifuged,  allowed  to
settle or filtered.

          7.3.8.1   Centrifugation:   Centrifugation at  2,000-3,000 rpm
      for 10 minutes is usually  sufficient to  clear the supernatant.
                           3015  -  7                        Revision  0
                                                          September 1994

-------
                  7.3.8.2   Settling:   Allow  the sample  to stand  until  the
               supernatant  is  clear.  Allowing  a sample to stand overnight will
               usually accomplish this.  If it does not, centrifuge or filter the
               sample.

                  7.3.8.3  Filtering: The filtering apparatus must be
               thoroughly cleaned  and prerinsed with dilute  (approximately 10%
               V/V)  nitric  acid.  Filter the sample through   qualitative filter
               paper into a second  acid-cleaned container.

               7.3.9  The concentration values  obtained  from analysis must be
         corrected for the dilution factor  from  the acid addition.  If the sample
         will  be  analyzed   by ICP-MS  additional  dilution  will  generally  be
         necessary.   For example,  the sample may  be diluted  by  a factor of 20
         with  reagent water and the acid strength  adjusted back  to 10% prior to
         analysis.   The  dilutions used should be recorded and the measured con-
         centrations adjusted  accordingly  (e.g., for a 45 ml  sample and 5 ml of
         acid  the correction factor is  1.11).

8.0  QUALITY CONTROL

         8.1   All  quality  control  measures described in Chapter  One,  of this
Manual, should be followed.

         8.2  For each analytical  batch of  samples  processed,  analytical reagent
blanks (also field  blanks  if  they were taken)  should be carried throughout the
entire sample preparation  and analytical process.   These blanks will be useful
in determining if samples  are being contaminated.

         8.3   Duplicate  samples  should be  processed on a  routine  basis.   A
duplicate sample is a real  sample brought through the whole  sample preparation
and  analytical  process.   A  duplicate  sample   should  be processed  with  each
analytical batch or every  20  samples, whichever is the greater number.

         8.4   Spiked samples or standard reference materials  should be employed
to determine accuracy.   A  spiked  sample should be included with each group of
samples processed and whenever a new sample matrix is being  analyzed.

9.0  METHOD PERFORMANCE

         9.1    Refer to  Table  1 for a summary of performance  data.
                                   3015 - 8                       Revision 0
                                                                  September 1994

-------
10.0  REFERENCES

1.       Introduction  to  Microwave Sample  Preparation:  Theory  and Practice.
         Kingston,  H.  M.;  Jassie,  L.  B., Eds.; ACS Professional Reference Book
         Series: American  Chemical  Society,  Washington, DC,  1988;  Ch 6  & 11.

2.       1985 Annual Book of ASTM Standards.  Vol.  11.01; "Standard Specification
         for Reagent Water";  ASTM:  Philadelphia,  PA,  1985;  D1193-77.

3.       Kingston,  H.  M.,  Final  Report EPA IAG #DWI3932541-01-I, September 30,
         1988, Appendix A.

4.       Shannon, M.,  Alternate  Test  Procedure Application,  USEPA  Region V,
         Central Regional  Laboratory,  536  S.  Clark  Street,  Chicago,  IL  60606,
         1989.

5.       Kingston,  H.  M.,  Walter,  P.  J.,  "Comparison   of Microwave  Versus
         Conventional Dissolution for Environmental  Applications", Spectroscopy,
         vol. 7 No. 9,20-27,1992.

6.       Sosinski,  P., and Sze C.,  "Absolute  Accuracy  Study, Microwave Digestion
         Method  3015  (Nitric acid  only)";   EPA  Region  III Central  Regional
         Laboratory, 1991.
                                   3015 -  9                       Revision 0
                                                                  September 1994

-------
                      TABLE 1
MICROWAVE DIGESTION METHOD  3015 (Nitric Acid Only)
Ele»
Al
Al
Al
Al
Ba
Ba
Ba
Cd
Cd
Cd
Cd
Zn
Zn
Zn
Zn
As
As
Co
Co
K
K
Ni
Ni
Ni
Pb
Pb
Pb
Pb
Sb
Sb
Se
Se
Tl
Tl
V
V
Be
Be
Ca
Ca
Material
Tm-11
Tm-12
T-107
T-109
Tm-11
Tm-12
T-107
Tm-11
Tm-12
T-107
T-109
Tm-11
Tm-12
T-107
T-109
T-107
T-109
Tm-11
Tm-12
T-95
T-109
Tm-11
Tm-12
T-109
Tm-11
Tm-12
T-107
T-109
UP980-1
WP980-2
T-95
T-107
UP980-1
WP980-2
Tm-11
Tm-12
T-107
T-109
T-107
T-109
Certified
Mean
510.0
2687.0
220.0
113.0
450.0
2529.0
192.0
40.8
237.0
14.3
12.1
55.4
314.0
75.8
74.0
10.8
8.15
227.0
1067.0
4700.0
2330.0
264.0
1234.0
57.0
275.0
1326.0
26.0
34.9
16.9
101.5
60.1
11.0
50.0
6.3
491.0
2319.0
11.0
22.1
11700.0
35400.0
Observed
Mean
485.5
2770.6
213.5
117.7
441.4
2431.4
196.6
44.6
242.3
12.4
10.3
55.9
316.5
81.6
69.9
12.8
90.6
242.6
1153.3
5080.3
2601.5
284.3
1293.0
60.8
275.9
1359.0
30.0
39.3
18.3
108.9
65.9
13.0
55.1
7.0
532.6
2412.8
11.3
25.6
12364.0
38885.0
Std. Dev.
26.3
88.2
19.3
30.6
23.4
70.3
15.9
2.1
8
0.9
1.7
2.6
8.9
3.3
4.1
0.84
11.0
14.1
35.9
784
383.4
16.5
39.4
3.09
32.2
35.0
0.2
1.2
0.47
34.4
2.6
0.9
2
0.52
26.1
60.6
0.53
0.91
783.6
999
Relative
Standard
Deviation
5.4
3.2
9.0
2.6
5.3
2.9
8.1
4.7
3.3
7.2
16.5
4.6
2.8
4.0
5.8
6.5
12.2
5.8
3.1
15.4
14.7
5.8
3.0
5.0
11.7
2.6
0.66
3.0
2.6
31.6
3.94
6.9
3.6
7.4
4.9
2.5
4.7
3.6
6.3
2.6
Relative
Bias
-4. 80%
3.11%
-2.95%
4.16%
-1.90%
-3.86%
2. 44%
9.46%
2.25%
-12.94%
-14.55%
1.06%
0.82%
7.68%
-5.46%
19.26%
11.26%
6.90%
8.09%
8.09%
11.65%
7.71%
4.79%
6.72%
0.36%
2.49%
15.65%
12.69%
8.27%
7.33%
9.77%
19.00%
10.26%
11.66%
8.48%
4.05%
3.00%
15.97%
5.68%
9.84%
                     3015 - 10
Revision 0
September 1994

-------
TABLE 1  (continued)
El em
Ca
Ca
Mg
Mg
Mg
Na
Na
Na
Cr
Cr
Cr
Cr
Cu
Cu
Cu
Cu
Fe
Fe
Fe
Fe
Mn
Mn
Mn
Mn
Ag
Material
T-107
T-109
T-95
T-107
T-109
T-95
T-107
T-109
Tm-11
Tm-12
T-107
T-109
Tm-11
Tm-12
T-107
T-109
Tm-11
Tm-12
T-107
T-109
Tm-11
Tm-12
T-107
T-109
WS378-1
Certified
Mean
11700.0
35400.0
32800.0
2100.0
9310.0
190000.0
20700.0
12000.0
52.1
299.0
13.0
18.7
46.3
288.0
30.0
21.4
249.0
1089.0
52.0
106.0
46.0
263.0
45.0
34.0
46.0
Observed
Mean
12364.0
38885.0
35002.0
2246.7
10221.7
218130.0
22528.0
13799.5
64.3
346.0
22.3
32.6
76.5
324.0
42.3
54.0
289.3
1182.5
63.8
134.0
60.9
304.4
52.6
46.6
19.4
Std. Dev.
783.6
999
1900
110.5
218.6
10700
1060
516.2
4.1
9.8
1.5
6.4
4.4
8.9
4.0
3.6
16.4
43.5
8.7
6.6
3.2
9.1
3.1
3.0
5.6
Relative
Standard
Deviation
6.3
2.6
5.4
4.9
2.1
4.9
4.7
3.7
6.4
2.8
6.7
19.6
5.7
2.7
9.4
6.7
5.7
3.7
13.6
4.9
5.2
3.0
5.9
6.4
2.9
Relative
Bias
5.68%
9.84%
6.71%
6.99%
9.79%
14.81%
8.83%
15.00%
23.51%
15.74%
71.77%
74.71%
65.36%
12.52%
41.17%
152.38%
16.18%
8.59%
22.69%
26.50%
32.48%
15.77%
17.09%
37.18%
-57.83%
     3015 - 11
Revision 0
September 1994

-------
                                       METHOD 3015
MICROWAVE  ASSISTED  ACID  DIGESTION OF  AQUEOUS SAMPLES  AND  EXTRACTS
              7.1 Calibrate
             the miorewavo
               equipment.
              7.2 AeM wee
              •nd HjO rlneo
              •II dioeotian
               OUoeware.
7.3.2 Moaeuro
46 mL aliquot
   Into the
  dlfeetlon
7.3.3 UM blank
  eamploe of
         p In
                                        7.3,4 Add
                                      eono*ntrat*
                                      HNOato
                                       7.3.( Plae*
                                       v*«««li in tn«
                                     blank* It n«e«**rv
                                     to
                                        7.3.6 Ploe*
                                        tho eorouool
                                       in ovon. hoot
                                        •eoording to
                                      powor program.
                                        7.3.7 Allow
                                        oool «o thoy
                                        •ra nat hot
                                         ta touch.
                                        7.3.1 r>looo
                                         ••mola in
                                        aeid-ol«anod
                                          bottlo.
                     7.S.*1-7.3.8.3
                       Contri«ugo.
                       aottlo, and
                      firtar aampla.
                                                              7.3.9 Carroet
                                                              oonoamration
                                                               valuoi for
                                                               tho dilution
                                                                 footer.
                                                                 I
                                        3015  -  12
                       Revision  0
                       September 1994

-------
                                 METHOD 3020A

                ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS
              FOR TOTAL METALS FOR ANALYSIS BY GFAA SPECTROSCOPY
1.0  SCOPE AND APPLICATION

      1.1   This digestion  procedure  is used  for  the preparation  of aqueous
samples, mobility-procedure extracts,  and wastes that contain suspended solids
for analysis  by  furnace  atomic absorption spectroscopy  (GFAA)  for  the metals
listed below.  The procedure is used to  determine the total amount of the metal
in the sample.

      1.2   Samples prepared  by Method  3020 may be  analyzed by GFAA for the
following metals:

            Beryllium                     Lead
            Cadmium                       Molybdenum
            Chromium                      Thallium
            Cobalt                        Vanadium

      NOTE; For the digestion  and  GFAA  analysis of arsenic  and selenium, see
            Methods 7060  and 7740.   For the  digestion  and GFAA analysis  of
            silver, see Method 7761.


2.0  SUMMARY OF METHOD

      2.1   A mixture  of nitric acid and the material to be analyzed is refluxed
in a covered Griffin beaker.  This  step  is repeated with additional  portions of
nitric  acid  until   the  digestate  is  light  in  color  or  until  its  color has
stabilized.  After the digestate has been brought to a low volume, it  is cooled
and brought up in  dilute  nitric  acid  such that the  final  dilution  contains 3%
(v/v) nitric  acid.  This  percentage will vary  depending  on  the amount of acid
used to complete the digestion.  If  the sample contains  suspended solids, it must
be centrifuged, filtered,  or allowed to settle.


3.0  INTERFERENCES

      3.1   Interferences are discussed in the referring analytical  method.


4.0  APPARATUS AND MATERIALS

      4.1   Griffin beakers - 150-mL,  or equivalent.

      4.2   Watch glasses - ribbed or  equivalent.


                                   3020A - 1                       Revision 1
                                                                  July 1992
'0

-------
      4.3   Qualitative filter paper or centrifugation equipment.

      4.4   Funnel or equivalent.

      4.5   Graduated Cylinder - lOOmL.

      4.6   Electric  hot  plate  or  equivalent  -  adjustable  and  capable  of
maintaining a temperature of 90-95°C.


5.0  REAGENTS

      5.1   Reagent grade chemicals  shall be used in all tests. Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such specifications are available. Other grades may be used,  provided it is first
ascertained that the  reagent  is  of  sufficiently high  purity to permit  its use
without lessening the accuracy of the determination.

      5.2   Reagent  Water.  Reagent  water will  be  interference  free.    All
references to  water  in  the method  refer  to reagent water  unless  otherwise
specified.  Refer to Chapter One for a definition of reagent water.

      5.3   Nitric  acid   (concentrated),  HN03.   Acid should  be  analyzed  to
determine levels of impurities.  If method blank is  < MDL, the  acid  can be used.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must   have been  collected  using  a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with detergents, acids, and
water.  Plastic and glass  containers are both suitable. See  Chapter Three, Step
3.1.3, for further  information.
      6.3   Aqueous wastewaters must be acidified to a pH of < 2 with HNO
                                                                         3'


7.0  PROCEDURE

      7.1   Transfer a 100-mL representative aliquot of the well-mixed sample to
a 150-mL Griffin beaker and add 3 mL of concentrated HN03.  Cover the  beaker with
a ribbed watch glass.  Place the beaker on  a hot plate and cautiously evaporate
to a low volume (5 mL),  making certain that the sample does not  boil  and that no
portion of the bottom of the  beaker is allowed to go dry.  Cool the beaker and
add another 3-mL portion of  concentrated HN03.  Cover  the  beaker with a non-
ribbed watch glass and return to  the hot plate.  Increase the  temperature of the
hot plate so that  a gentle reflux action occurs.


                                   3020A -  2                       Revision 1
                                                                  July 1992

-------
      7.2   Continue heating,  adding  additional  acid as  necessary,  until  the
digestion is complete (generally indicated  when the digestate is light in color
or does not change  in appearance with continued refluxing). When the  digestion
is complete, evaporate  to  a low volume (3 ml); use  a  ribbed  watch glass,  not
allowing any portion of the bottom of the beaker to go dry.  Remove the beaker
and add approximately 10 ml of water, mix,  and continue warming the beaker for
10 to 15 minutes to allow additional solubilization of any residue to occur.

      7.3   Remove the beaker from the hot  plate and  wash  down the beaker walls
and watch glass with water.   When  necessary, filter or centrifuge the sample to
remove silicates and other  insoluble material  that may interfere with injecting
the sample into the graphite atomizer.  (This additional step can cause sample
contamination unless the filter and filtering apparatus are thoroughly cleaned
and prerinsed  with dilute  HN03.)   Adjust  to  the  final  volume of  100  ml with
water.  The sample is now ready for analysis.


8.0  QUALITY CONTROL

      8.1   All quality  control  measures  described  in  Chapter One  should be
followed.

      8.2   For each batch of samples processed, method blanks should be carried
throughout the entire sample preparation and analytical process.   These blanks
will   be  useful in  determining  if samples are  being contaminated.   Refer to
Chapter One for the proper protocol when analyzing blanks.

      8.3   Replicate samples should be processed on a routine basis.  Replicate
samples will  be used to determine precision.  The  sample load  will  dictate
frequency, but 5% is recommended.   Refer to Chapter One for the proper protocol
when analyzing replicates.

      8.4   Spiked samples  or standard reference materials should be employed to
determine  accuracy.   A  spiked  sample should be  included with each  batch of
samples processed  or  5% and whenever  a  new sample  matrix is  being  analyzed.
Refer to Chapter One for the proper protocol  when analyzing spikes.

      8.5   The concentration of  all  calibration  standards should be verified
against a quality control check sample obtained from an outside source.  Refer
to Chapter One for the proper protocol.

      8.6   The method of standard addition shall be used for the analysis of all
EP extracts.  See Method 7000, Step 8.7,  for further information.


9.0  METHOD PERFORMANCE

      9.1   No data provided.
                                   3020A  -  3                       Revision 1
                                                                  July 1992

-------
10.0  REFERENCES

1.    Rohrbough, W.G.;  et  al.  Reagent Chemicals,  American Chemical  Society
Specifications, 7th ed.; American Chemical Society:  Washington,  DC, 1986.

2.    1985 Annual Book of ASTM Standards, Vol. 11.01;  "Standard Specification for
Reagent Water"; ASTM: Philadelphia, PA, 1985; D1193-77.
                                   3020A - 4                      Revision 1
                                                                  July  1992

-------
                       METHOD  3020A
  ACID  DIGESTION FOR AQUEOUS  SAMPLES AND EXTRACTS
FOR  TOTAL METALS  FOR ANALYSIS BY GFAA SPECTROSCOPY
                          Start
                      7.1 Put sample
                       aliquot in
                       beaker, add
                      concentrated
                     HNO,,evaporate to
                       low volume
                    7.1 Cool beaker,add
                      concentrated
                     HNO,,heat until
                      gentle raflux
                      action occur*
                       7 2  Haat to
                    complete digestion,
                    evaporata to low
                       volume,cool
                     7 2 Add reagent
                      watar,warm to
                      dissolve any
                     precipitate or
                         re»idue
                      7 3 Filter or
                      centrifuge if
                      necessary and
                      adjust volume
                         3020A -  5
Revision 1
July 1992

-------
                                 METHOD 3040

              DISSOLUTION PROCEDURE FOR OILS.  GREASES.  OR WAXES
1.0  SCOPE AND APPLICATION

     1.1  Method 3040 is used for  the preparation of samples containing oils,
greases, or waxes  for  analysis  by  atomic  absorption spectroscopy (AAS) or
inductively coupjed argon plasma emission spectroscopy (ICP) for the following
metals:
Antimony
Beryl 1 i urn
Cadmium
Chromium
Copper
                                             Iron
                                             Manganese
                                             Nickel
                                             Vanadium
     1.2  This method is  a  solvent  dissolution  procedure,  not a digestion
procedure.  This procedure can be  very  useful  in the analysis of crude oil,
but with spent or used oil high  in parti cul ate material it is less effective;
most parti cul ate material is not dissolved,  and therefore the analysis is not
a "total" metal determination.   Because  the  highest percentage of metals is
expected to be  contained  in  the  particulate  material,  oil analysis using
Method 3040  will  not  provide  an  adequate  estimate  of  the  total metals
concentration.
2.0  SUMMARY OF METHOD

     2.1  A representative  sample  is  dissolved  in  an  appropriate solvent
(e.g., xylene  or  methyl   isobutyl  ketone).    Organometal11c  standards are
prepared using the same solvent, and the samples and standards are analyzed by
AAS or ICP.

3.0  INTERFERENCES

     3.1  Diluted  samples  and  diluted  organometallic  standards  are often
unstable.  Once standards and  samples  are diluted, they should be analyzed as
soon as possible.

     3.2  Solvent  blanks   should  be  used  to  rinse  nebulizers  thoroughly
following aspiration  of high concentration standards or samples.

     3.3  Viscosity   differences   can  result  in  different  rates  of sample
introduction;  therefore,  all   analyses  shall  be  performed  by the method of
standard addition.    Peristaltic   pumps  often   prove  useful when analysis is
performed by  ICP.
                                   3040 - 1
                                                          Revision       0
                                                          Date   September  1986

-------
4.0  APPARATUS AND MATERIALS

     4.1  Volumetric glassware.

     4.2  Balance.

     4.3  Atomic absorption spectrometer;   With  an auxiliary oxldant control
and a mechanism for background correction.

     4.4  Inductively coupled argon plasma emission spectrometer system;  With
a  mechanism?orbackground   correctionandInterelementinterference
correction.  A peristaltic pump 1s optional.


5.0  REAGENTS

     5.1  Methyl Isobutyl ketone (MIBK).

     5.2  Xylene.

     5.3  Organometal11c  standards    (two   possible   sources  are  Conostan
Division, Conoco SpecialityProducts,   Inc.,  P.O.  Box  1267, Ponca City, OK
74601, and the  U.S.  Department  of   Commerce,  National Bureau of Standards,
Washington, DC 20234).


6.0  SAMPLE COLLECTION,  PRESERVATION,  AND HANDLING

     6.1  All samples must   have  been  collected   using  a sampling plan that
addresses the considerations discussed 1n Chapter Nine of this manual.

     6.2  Samples  shall  be  stored  in  an undiluted state at room temperature.

     6.3  Samples  should be processed and analyzed  as soon as possible.


7.0  PROCEDURE

     7.1  Weigh  out a   2-g   representative   sample   of  the  waste  or extract.
Separate  and  weigh the  phases  if more than  one phase is present.

     7.2  Weigh  an aliquot  of  the  organic  phase and dilute the aliquot in  the
appropriate solvent. Warming   facilitates   the subsampling of crude-type  oils
and  greases and wax-type wastes.    Xylene is usually the  preferred  solvent  for
longer-chain  hydrocarbons and  for  most analyses performed by  ICP.   The  longer-
chain  hydrocarbons usually require a  minimum  of a 1:10  dilution,  and  lighter
oils may  require only  a 1:5 dilution if low detection limits  are  required.

      7.3   All  metals must be  analyzed  by   the  method  of  standard additions.
Because the method of standard  additions  can account only for multiplicative
 interferences (matrix  or physical   interferences),  the analytical program  must


                                   3040 - 2
                                                          Revision       0
                                                          Date  September 1986

-------
account for additive Interference (nonspecific absorption and scattering 1n
AAS  and  nonspecific  emission  and  Interelement  Interference  1n  ICP)  by
employing background correction.

     7.4  Sample preparation  for  the  method  of  standard  additions can be
performed on a weight  or  volume  basis.    Sample allquots of viscous wastes
should be weighed.  Weigh  Identical  amounts  of  the sample Into three wide-
mouth vials.  Dilute the first vial such that the final concentration falls on
the lower end of the linear portion of the calibration curve and significantly
above the detection limit.  Add  sufficient  standard to the second aliquot to
Increase the sample  concentration  by  approximately  50%.   Adjust the third
sample concentration so that 1t 1s approximately twice that of the first.  The
second and third allquots are  then  diluted  to  the same final volume as the
first aliquot.

     7.5  Set up and calibrate the analytical Instrumentation according to the
manufacturer's directions for nonaqueous samples.

     7.6  Report data as the weighted average for all sample phases.


8.0  QUALITY CONTROL

     8.1  Preparation blanks  (e.g.,  Conostan  base  oil  or  mineral oil plus
reagents)   should   be  carried  through  the  complete   sample-preparation and
analytical  process  on  a  routine  basis.    These  blanks  will  be useful In
detecting and determining the magnitude of any sample contamination.

     8.2  Duplicate samples should be processed on a routine basis.  Duplicate
samples will be used to determine precision.  The sample load will dictate the
frequency,  but 20%  1s recommended.

     8.3  Samples and standards should  be  diluted  as  closely as possible to
the time of analysis.

     8.4  All analyses must be performed  by the method  of standard additions.
See Method  7000, Section 8.7, for further Information.

     8.5  Data must be corrected  for  background  absorption and emission and
Interelement Interferences.


9.0  METHOD PERFORMANCE

     9.1  No data provided.


10.0   REFERENCES

     10.1   None required.
                                   3040 - 3
                                                          Revision       0
                                                          Date  September  1986

-------
                                        METHOD 3040

                     DISSOLUTION PROCEDURE FOR OILS.  GREASE.  OR  MAXES
7.1
                                                                               o
  Weigh out
    2 g of
    •ample
                                                                          7.4
                                                       Weigh  sample
                                                      into  3  vials:
                                                 dilute 1st vial:  add
                                                 standard to  2nd vial
                                                 to  increase  cone. by
                                                 SOX;  adjust  3rd vial
                                                  cone,  to  twice the
                                                cone,  of the  1st vial
Is more than
  one phase
  present?
Separate and
weigh phases
7.4
    I  Dilute
    second and
third  aliquots
to same volume
    as first
7.2
J       Weigh
      aliquot
    of organic
 phase:  dilute
 with appropr.
     solvent
                                                          Set up
                                                     and calibrate
                                                       analytical
                                                   instrumentation
7.3
    I  Analyre
     octal* by
     standard
     additions
      •method
                                                                             7.6
                                                   Report data as
                                                     weighted
                                                     everege
                                                                           f     Stop      J
                                   3040 - 4
                                                              Revision       0
                                                              Date  September  1986

-------
                                 METHOD 3050A

                ACID  DIGESTION OF  SEDIMENTS,  SLUDGES, AND  SOILS
1.0   SCOPE AND APPLICATION

      1.1   This method is an acid digestion procedure used to prepare sediments,
sludges, and  soil  samples  for analysis by flame  or  furnace atomic absorption
spectroscopy (FLAA and GFAA, respectively)  or by inductively coupled argon plasma
spectroscopy (ICP).  Samples prepared by this  method may  be  analyzed by ICP for
all the listed metals, or by FLAA or GFAA as indicated below (see also Step 2.1):

            	FLAA	                 GFAA

            Aluminum                Magnesium               Arsenic
            Barium                  Manganese               Beryllium
            Beryllium               Molybdenum              Cadmium
            Cadmium                 Nickel                  Chromium
            Calcium                 Osmium                  Cobalt
            Chromium                Potassium               Iron
            Cobalt                  Silver                  Lead
            Copper                  Sodium                  Molybdenum
            Iron                    Thallium                Selenium
            Lead                    Vanadium                Thallium
                                    Zinc                    Vanadium

      NOTE: See Method 7760 for FLAA preparation for Silver.


2.0  SUMMARY OF METHOD

      2.1  A representative 1- to 2-g  (wet weight) sample is digested in nitric
acid and hydrogen peroxide.  The digestate is then refluxed with either nitric
acid or hydrochloric acid.  Hydrochloric  acid  is used  for flame AA  and ICP
analyses and nitric acid is used for furnace AA work.  Dilute hydrochloric acid
is used as the final  reflux acid for (1) the ICP analysis of As  and Se, and (2)
 he flame AA or ICP analysis of Ag, Al, Ba,  Be,  Ca,  Cd, Co, Cr, Cu, Fe, K, Mg,
    Mo, Na, Ni, Os,  Pb, Tl, V,  and Zn.  Dilute  nitric acid is employed as the
final dilution acid for the furnace AA  analysis of As, Be, Cd, Cr, Co, Fe, Pb,
Mo, Se, Tl, and V.  The diluted samples have an approximate acid concentration
of  5.0%  (v/v).    A  separate  sample  shall  be  dried for  a  total  %  solids
determination.
3.0  INTERFERENCES

      3.1   Sludge samples can contain diverse matrix types, each of which may
present its own analytical  challenge.   Spiked  samples and any relevant standard
reference material should be processed to aid in determining  whether Method 3050
is applicable to a given waste.
                                   3050A  -  1                      Revision 1
                                                                 July 1992

-------
4.0  APPARATUS AND MATERIALS

      4.1   Conical Phillips beakers - 250-mL,  or equivalent.

      4.2   Watch glasses ribbed or equivalent.

      4.3   Drying ovens - That can be maintained at 30°  C.

      4.4   Thermometer - That covers range of 0-200°C.

      4.5   Filter paper - Whatman No. 41 or equivalent.

      4.6   Centrifuge and centrifuge tubes.

      4.7   Analytical Balance - Capable of accurately weighing to the
nearest 0.01 g.

      4.8   Electric  Hot Plate  or  equivalent  -  Adjustable  and capable  of
maintaining a temperature of 90-95°C.

      4.9   Glass Funnel or equivalent.

      4.10  Graduated cylinder or equivalent.


5.0  REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all  reagents shall conform to the specifications
of the Committee on Analytical  Reagents of  the  American Chemical Society, where
such specifications are  available. Other grades  may  be used,  provided it  is first
ascertained that  the  reagent  is  of sufficiently high  purity to permit its use
without lessening the accuracy of the determination.  If the  purity of a reagent
is questionable, analyze the reagent to determine the  level  of impurities.  The
reagent blank must be less than the MDL in order to be used.

      5.2   Reagent  Water.  Reagent  water  will be  interference  free.    All
references  to water  in the  method  refer  to  reagent water  unless  otherwise
specified.  Refer to Chapter One for a definition of reagent water.

      5.3   Nitric  acid  (concentrated),  HN03.   Acid should  be  analyzed  to
determine level of impurities.  If method blank is  < MDL, the acid  can  be used.

      5.4   Hydrochloric acid  (concentrated),  HC1.  Acid should be analyzed to
determine level of impurities.  If method blank is  < MDL, the acid  can  be used.

      5.4   Hydrogen  peroxide  (30%),  H202.    Oxidant should  be  analyzed  to
determine level of impurities.
                                   3050A - 2                     Revision 1
                                                                 July 1992

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must have been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with detergents, acids, and
water.  Plastic and  glass containers are both suitable.  See  Chapter Three, Step
3.1.3, for further  information.

      6.3   Nonaqueous samples shall  be refrigerated upon receipt and analyzed
as soon as possible.


7.0  PROCEDURE

      7.1   Mix the  sample thoroughly to achieve homogeneity.  For each digestion
procedure, weigh to  the nearest 0.01 g and transfer to a conical beaker 1.00-2.00
g of sample.   For samples with  low percent  solids  a larger sample size may be
used as long as digestion is completed.

      7.2   Add 10 mL of 1:1  HN03, mix the slurry,  and cover  with a watch glass.
Heat the sample to 95°C  and  reflux for  10 to  15 minutes without boiling.  Allow
the sample to cool,  add 5 mL of concentrated  HN03, replace the watch glass, and
reflux for  30  minutes.   Repeat  this last step to  ensure  complete oxidation.
Using a  ribbed watch glass, allow the solution  to  evaporate  to  5 mL without
boiling, while maintaining  a covering of  solution over the bottom of the beaker.

      7.3   After Step 7.2 has been completed and the  sample has cooled, add 2 mL
of water and 3 mL of  30% H202.   Cover the beaker with a watch glass and return
the covered  beaker to  the  hot  plate  for warming  and  to  start  the peroxide
reaction.   Care  must be  taken  to  ensure  that  losses do  not occur  due  to
excessively vigorous effervescence.  Heat until effervescence  subsides and cool
the beaker.

      7.4   Continue  to  add  30%  H202 in  1-mL aliquots  with warming  until  the
effervescence  is minimal or until the general sample appearance is unchanged.

      NOTE: Do not  add more than  a total  of 10 mL 30% H202.

      7.5   If the  sample is being prepared for (a)  the ICP analysis of As and
Se, or (b) the flame AA or ICP analysis  of Ag, Al, Ba, Be,  Ca, Cd, Co, Cr, Cu,
Fe, K, Mg, Mn,  Mo, Na, Ni,  Os,  Pb, Tl,  V, and  Zn, then add 5 mL of concentrated
HC1 and 10 mL  of water, return the covered beaker to the hot plate, and reflux
for an additional 15 minutes  without  boiling.  After cooling, dilute to a  100 mL
volume with water.   Particulates in  the digestate  that may clog the nebulizer
should be removed by filtration,  by centrifugation, or by allowing  the sample to
settle.

            7.5.1    Filtration  -  Filter through Whatman No.  41 filter paper (or
      equivalent).
                                   3050A -  3                     Revision 1
                                                                 July 1992

-------
            7.5.2   Centrifugation  -  Centrifugation  at  2,000-3,000 rpm  for
      10 minutes is usually sufficient to clear the supernatant.

            7.5.3   The diluted sample has an approximate  acid concentration of
      5.0% (v/v) HC1  and  5.0%  (v/v) HN03.  The sample is now ready for analysis.

      7.6   If the sample is being prepared  for the furnace analysis of As, Be,
Cd, Co,  Cr,  Fe, Mo, Pb, Se, Tl, and V, cover the sample  with a ribbed watch glass
and continue  heating the  acid-peroxide digestate until  the  volume  has  been
reduced to  approximately  5 ml.   After  cooling,  dilute to 100  ml  with  water.
Particulates  in  the  digestate   should  then  be  removed by   filtration,  by
Centrifugation, or by allowing the sample to settle.

            7.6.1   Filtration -  Filter through Whatman No. 41 filter paper (or
      equivalent).

            7.6.2   Centrifugation  -  Centrifugation  at  2,000-3,000 rpm  for
      10 minutes is usually sufficient to clear the supernatant.

            7.6.3   The  diluted  digestate  solution contains  approximately 5%
      (v/v) HN03.   For analysis, withdraw aliquots of appropriate volume and add
      any required reagent or matrix modifier.   The  sample is now  ready for
      analysis.

      7.7   Calculations

            7.7.1   The  concentrations determined are to be reported  on the
      basis of  the actual  weight  of the sample.   If  a dry  weight  analysis is
      desired, then the percent solids of the sample must also be provided.

            7.7.2    If percent solids  is desired, a  separate determination of
      percent solids must  be  performed on a homogeneous aliquot of the sample.


8.0  QUALITY CONTROL

      8.1   All  quality  control   measures described  in Chapter One  should be
followed.

      8.2   For each batch of samples  processed,  preparation  blanks should be
carried throughout the entire sample preparation  and analytical  process.  These
blanks will be useful in determining if samples are being contaminated.  Refer
to Chapter One for the proper protocol when analyzing blanks.

      8.3   Replicate samples should be processed  on a routine  basis.  Replicate
samples will  be used to  determine precision.   The  sample load  will  dictate
frequency, but 5% is recommended.   Refer to  Chapter One for the  proper protocol
when analyzing replicates.

      8.4   Spiked samples  or standard reference  materials must be employed to
determine  accuracy.   A  spiked sample should  be  included with each batch of
                                   3050A - 4                     Revision 1
                                                                 July  1992

-------
samples processed and whenever a new sample matrix is being analyzed.  Refer to
Chapter One for the proper protocol when analyzing spikes.

      8.5   The concentration of all calibration  standards  should be verified
against a quality control check sample obtained from an outside source.


9.0  METHOD PERFORMANCE

      9.1   No data provided.


10.0  REFERENCES

1.    Rohrbough,  W.G.;  et  al.  Reagent  Chemicals,  American  Chemical  Society
Specifications, 7th ed.; American Chemical Society:  Washington, DC, 1986.

2.    1985 Annual  Book of ASTM Standards, Vol. 11.01;  "Standard  Specification for
Reagent Water"; ASTM: Philadelphia, PA, 1985; D1193-77.

3.    Edgell, K.; USEPA Method Study 37 - SW-846 Method 3050 Acid Digestion of
Sediments, Sludges, and Soils.  EPA Contract No. 68-03-3254, November 1988.
                                  3050A  -  5                      Revision 1
                                                                 July 1992

-------
                         METHOD 3050A
  ACID  DIGESTION  OF  SEDIMENTS,  SLUDGES,  AND SOILS
                             Start
            Furnace
            analysis
            for As, Be,
            Cd ,  Co , Cr ,
            Fe,  Mo, Pb,
            Se,  Tl, V
 7.6 Continue
  heating to
 reduce volume
7.6 Dilute with
 reagent  water
  and  filter
particulates in
   digastate
                            7.1 Mix
                         sample; take
                         1-2 g portion
                           for each
                           diges tion
                          7 2 Add HNO,,
                          reflux;repeat
                          HNO, raflux
                        until solution
                            i* 5 ml
                         7.3 Add raagant
                         watar and H,0t;
                         heat baakar to
                         start peroxide
                           reaction
                         7 4 Continue
                          adding H,0,
                         with heating
 7.7.1  Report
concentrations,
and % solids of
sample  for dry
weight  analysis
                ICP or  Flame AA
                analysis for
                As,  Ag, Al  Ba,
                Be,  Ca, Cd
                    Cu, Fe
                    Mn, Mo
                Ni,  0., Pb
                Tl,  V,  Zn
   Co,
   K,
   Na,
   Sa,
                            7.S Add
                         concentrated
                            HC1 and
                            reagent
                         water; reflux
7.5 Cool;diluta
 with reagent
 water,  filter
partculates in
   digastate
                           7.7.2 If %
                            solids
                         required,use
                           homogeneous
                         sample aliquot
                             Stop
                            3050A -  6
                                            Revision  1
                                            July  1992

-------
                                  METHOD 3051

               MICROWAVE ASSISTED ACID DIGESTION OF SEDIMENTS,
                           SLUDGES, SOILS, AND OILS
 1.0   SCOPE AND APPLICATION

      1.1  This method is  applicable to the microwave assisted acid  digestion of
 sludges, sediments, soils, and oils for the following elements:

      Aluminum    Cadmium         Iron          Molybdenum          Sodium
      Antimony    Calcium         Lead          Nickel              Strontium
      Arsenic     Chromium        Magnesium     Potassium           Thallium
      Boron       Cobalt          Manganese     Selenium            Vanadium
      Barium      Copper          Mercury       Silver              Zinc
      Beryllium

      1.2   This method is provided as  an alternative to Method 3050.   It is
 intended to provide a rapid multielement acid  leach  digestion  prior to analysis
 so that  decisions can  be  made about site  cleanup  levels,  the need for TCLP
 testing  of  a  waste   and  whether  a  BOAT  process  is  providing  acceptable
 performance.   If  a decomposition including hydrochloric acid is  required for
 certain elements,  it is recommended that Method 3050A be used.  Digests produced
 by the  method are  suitable  for  analysis  by  flame  atomic  absorption (FLAA),
 graphite furnace atomic absorption (GFAA), inductively coupled plasma emission
 spectroscopy (ICP-ES)  and  inductively coupled plasma  mass spectrometry (ICP-MS).
 Due to the rapid advances in microwave technology, consult your manufacturer's
 recommended instructions  for guidance on  their  microwave  digestion  system and
 refer to the SW-846 "DISCLAIMER"  when conducting analyses using Method 3051.

 2.0  SUMMARY OF METHOD

      2.1   A representative sample  of  up to  0.5 g  is digested   in  10  mL of
 concentrated nitric  acid  for 10  min  using microwave heating with a suitable
 laboratory microwave unit.  The sample and  acid are placed in  a  fluorocarbon (PFA
 or TFM) microwave vessel.  The vessel  is  capped and heated in  the microwave unit.
 After cooling,  the  vessel contents are filtered,  centrifuged, or allowed to
 settle and then diluted to volume  and analyzed by the  appropriate SW-846 method
 (Ref. 1).

 3.0  INTERFERENCES

      3.1  Very reactive  or volatile materials  that may create high pressures
when heated may cause venting of the vessels with potential  loss of  sample and
 analytes.   The  complete decomposition  of  either carbonates, or carbon  based
 samples,  may cause  enough pressure to  vent the  vessel  if  the  sample  size is
greater than 0.25 g when used in the 120 mL vessels with a pressure relief device
that has  an upper limit of 7.5+ 0.7 atm (110 ± 10 psi).
                                   3051 - 1                       Revision 0
                                                                  September 1994

-------
4.0  APPARATUS AND MATERIALS

      4.1  Microwave apparatus requirements.

            4.1.1   The  microwave  unit provides  programmable  power with  a
      minimum of 574 W, which  can be programmed to within ±  10 W of the required
      power.   Typical units  provide  a nominal  600 W  to 1200  W  of  power.
      Pressure,  or especially  temperature,  monitoring  and  control  of  the
      microwave unit are desirable.

            4.1.2   The microwave  unit cavity is corrosion  resistant and well
      ventilated.

            4.1.3  All electronics are protected against corrosion for safe
      operation.

            4.1.4   The system  requires fluorocarbon  (PFA or TFM)  digestion
      vessels  (120 ml capacity) capable  of  withstanding  pressures  up to  7.5 ±
      0.7 atm  (110 ±  10  psi)  and  capable  of  controlled  pressure  relief at
      pressures exceeding 7.5 ± 0.7 atm (110 ± 10 psi).

            4.1.5  A  rotating  turntable  is  employed  to  insure  homogeneous
      distribution of microwave radiation within the unit.  The  speed  of the
      turntable should be a minimum of 3 rpm.

                  CAUTION:  Those  laboratories  now  using or contemplating the
                  use of kitchen type  microwave  ovens  for this method should be
                  aware of several  signifant safety issues.   First, when an acid
                  such as  nitric is used to assist sample  digestion in microwave
                  units in open  vessels,  or  sealed vesselsequippedres, there is
                  the potential  for  the  acid gases  released  to  corrode  the
                  safety  devices  that  prevent   the  microwave  magnetron  from
                  shutting off  when  the door is opened.  This can  result in
                  operator exposure to  microwave energy.  Use of  a  unit with
                  corrosion   resistant   safety   devices  prevents   this   from
                  occurring.

                  CAUTION:  The second safety  concern  relates  to the  use of
                  sealed containers without  pressure relief valves in the unit.
                  Temperature  is  the   important   variable   controlling  the
                  reaction.  Pressure  is needed to attain elevated temperatures
                  but must be  safely contained.  However,  many digestion vessels
                  constructed from certain  fluorocarbons may crack,  burst, or
                  explode in  the unit  under certain pressures.    Only unlined
                  fluorocarbon  (PFA  or TFM)  containers with pressure  relief
                  mecahnisms  or containers with PFA-fluorocarbon liners  and
                  pressure  relief  mechanisms  are   considered  acceptable  at
                  present.

                  Users are therefore  advised  not to use  kitchen type microwave
                  ovens or  to use  sealed  containers without pressure  relief

                                   3051 - 2                       Revision 0
                                                                  September 1994

-------
                  valves for microwave acid digestions by this method.  Use of
                  laboratory-grade microwave  equipment  is  required to prevent
                  safety hazards.  For further details consult reference 2.

                  CAUTION:      There   are   many    safety   and   operational
                  recommendations specific to the model  and manufacturer of the
                  microwave equipment  used  in individual  laboratories.  These
                  specific suggestions are beyond the scope of this method and
                  require the analyst to consult  the  specific equipment manual,
                  manufacturer and literature for proper and safe operation of
                  the microwave equipment and vessels.

      4.2   Volumetric graduated cylinder,  50 or  100  ml capacity or equivalent.

      4.3   Filter paper, qualitative or equivalent.

      4.4  Filter funnel, glass or disposable polypropylene.

      4.5  Analytical balance, 300 g capacity, and minimum ± 0.01 g.

5.0  REAGENTS

      5.1  All acids should be sub-boiling  distilled  where possible to minimize
the  blank  levels due to  metallic contamination.   Other  grades may  be  used,
provided it is  first ascertained that the reagent is of  sufficient purity to
permit  its  use  without  lessening the  accuracy  of  the determination.   If the
purity of a reagent is questionable,  analyze  the  reagent to determine the level
of impurities.  The reagent  blank  must be less than  the MDL  in order to be used.

            5.1.1  Concentrated nitric acid, HN03.   Acid should be analyzed to
      determine levels of impurity.   If the method blank is less than the MDL,
      the acid can be used.

      5.2   Reagent  Water.   Reagent water shall be interference  free.   All
references  to water  in  the method  refer to  reagent water unless  otherwise
specified (Ref.  3).

6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   All  samples must  have been collected  using a sampling  plan  that
addresses the considerations discussed in Chapter Nine of this  manual.

      6.2  All sample containers  must  be  prewashed  with detergents,  acids and
water.  Plastic  and  glass containers are both suitable.   See Chapter Three, sec.
3.1.3 of this manual, for further information.

     6.3  Samples  must  be refrigerated  upon  receipt and analyzed  as  soon as
possible.
                                   3051 - 3                       Revision 0
                                                                  September 1994

-------
7.0  PROCEDURE

      7.1  Calibration of Microwave Equipment

            NOTE:  If the  microwave  unit  uses  temperature  feedback  control
            capable of replicating the performance specifications of the method,
            then the calibration procedure may be omitted.

            7.1.1  Measurement of the available power for heating is evaluated
      so that absolute power in watts may be transferred from one microwave unit
      to another.  For cavity  type microwave equipment, this is accomplished by
      measuring  the  temperature rise  in  1 kg of  water  exposed  to  microwave
      radiation  for  a fixed period  of time.  The analyst can  relate  power in
      watts to the partial power setting of the  unit.  The  calibration format
      required for laboratory  microwave units depends on the type of electronic
      system used by the manufacturer to provide partial  microwave power.  Few
      units have  an  accurate  and precise  linear relationship  between  percent
      power  settings and  absorbed  power.    Where linear  circuits have  been
      utilized,  the  calibration  curve  can be  determined  by a  three-point
      calibration method (7.1.3), otherwise, the analyst  must  use the multiple
      point calibration  method (7.1.2).

            7.1.2  The multiple  point  calibration  involves  the measurement of
      absorbed power over a large range of power settings.   Typically,  for a
      600 W unit, the following  power  settings are measured;  100,  99,  98,  97,
      95, 90, 80, 70, 60, 50,  and 40% using the procedure described in section
      7.1.4.  This data  is clustered about the customary  working power ranges.
      Nonlinearity  has  been  commonly  encountered  at the  upper  end  of  the
      calibration.   If   the  unit's  electronics  are known  to have  nonlinear
      deviations  in  any  region of  proportional power  control,  it  will  be
      necessary to make  a set of measurements that bracket the power to be used.
      The final  calibration point should be at the  partial  power  setting that
      will be used in the test.   This  setting should be checked periodically to
      evaluate the integrity  of the calibration.  If a  significant  change is
      detected (±10 W),  then the entire calibration should  be  reevaluated.

            7.1.3   The   three-point  calibration involves  the measurement  of
      absorbed power at  three different  power  settings.  Measure  the  power at
      100% and 50% using athe  procedure described in section 7.1.4.   From the
      2-point  line calculate  the power setting  corresponding  to  the  required
      power in watts specified in the procedure.  Measure the absorbed power at
      that  partial  power setting.    If the measured  absorbed power does  not
      correspond to  the  specified  power within ±10 W, use  the multiple  point
      calibration in  7.1.2.   This point  should  also be used  to  periodically
      verify the integrity of  the calibration.

            7.1.4  Equilibrate a large volume of water to room temperature
      (23 ± 2°C).  One kg of reagent water is weighed (1,000.0 g ± 0.1  g) into
      a fluorocarbon beaker or  a beaker made of  some  other  material  that does
      not significantly  absorb microwave energy (glass absorbs microwave energy
      and is not recommended).  The initial temperature of the water should be

                                   3051 - 4                       Revision 0
                                                                  September  1994

-------
23  ±  2°C measured  to  ± 0.05°C.    The  covered  beaker  is  circulated
continuously  (in the normal  sample  path)  through the  microwave field for
2 minutes at the desired partial power setting with the unit's exhaust fan
on maximum (as it will  be during normal operation).  The beaker is removed
and  the  water vigorously stirred.   Use a magnetic stirring  bar inserted
immediately after microwave irradiation and record the maximum temperature
within  the first 30  seconds to ±  0.05°C.   Use  a new  sample  for  each
additional measurement.   If the water is  reused  both the water and the
beaker must have returned  to 23 ± 2°C.   Three measurements  at each power
setting  should  be made.

The  absorbed  power is  determined by the  following relationship:

                        P = (K) (Cp) (m)  (AT)
Eq.  1                  	
Where:

P = the apparent power absorbed by the  sample  in watts  (W)
(W=joule-sec~1)

K  =  the  conversion  factor for  thermochemical  calories-sec"1  to  watts
(=4.184)

Cp = the heat capacity, thermal capacity, or specific heat
(cal-g'1 "C"1)  of water

m = the mass of the water sample  in grams  (g)

AT  =  the  final temperature  minus  the  initial  temperature (°C)

t = the time in seconds  (s)


Using the experimental conditions  of 2 minutes  and 1 kg of distilled water
(heat capacity  at  25  °C  is 0.9997 cal-g~1-°C"1)  the calibration  equation
simplifies to:

Eq. 2                    P = (AT)  (34.86)

      NOTE:    Stable line voltage   is  necessary  for  accurate   and
      reproducible calibration and operation.   The line  voltage  should be
      within  manufacturer's specification,  and during  measurement  and
      operation should  not  vary  by  more than  ±2  V.    A constant  power
      supply may be necessary  for microwave use if the  source of the
      line voltage is unstable.

      Electronic components  in most  microwave  units  are matched to  the
      units'  function and  output.   When any  part of  the  high  voltage


                             3051 - 5                       Revision  0
                                                            September 1994

-------
            circuit, power source, or control components in the unit have been
            serviced or  replaced,  it will be necessary  to  recheck the units'
            calibration.  If the power output has changed  significantly  (±10 W),
            then the entire calibration should be reevaluated.

      7.2   All  digestion vessels  and  volumetric ware must  be  carefully acid
washed and rinsed with reagent water.  When switching between high concentration
samples and low concentration samples, all digestion vessels should be cleaned
by leaching with hot (1:1) hydrochloric acid  (greater than 80°C, but less than
boiling) for a minimum of two hours followed with hot (1:1) nitric acid  (greater
than 80°C, but less  than boiling)  for a minimum of  two  hours  and  rinsed with
reagent water and dried  in a clean  environment.   This cleaning procedure should
also be used whenever the prior use of the digestion vessels  is unknown  or cross
contamination from vessels is suspected.  Polymeric or glass  volumetric  ware and
storage  containers  should  be  cleaned  by  leaching  with  more dilute  acids
(approximately 10% V/V)  appropriate for  the specific plastics used  and then
rinsed  with  reagent water  and  dried  in  a  clean  environment.    To  avoid
precipitation of silver, ensure that all HC1 has been rinsed from the  vessels.

      7.3  Sample Digestion

            7.3.1  Weigh the fluorocarbon  (PFA or TFM) digestion vessel, valve
      and capassembly to 0.001 g prior to  use.

            7.3.2   Weigh a  well-mixed sample to the  nearest  0.001  g  into the
      fluorocarbon  sample  vessel  equipped  with  a  single-ported  cap  and  a
      pressure relief valve. For soils, sediments, and sludges use  no more than
      0.500 g.  For oils use no more than  0.250 g.

            7.3.3   Add  10  + 0.1  ml concentrated nitric  acid  in a  fume hood.
      If a vigorous reaction occurs, allow the reaction to stop before capping
      the vessel.   Cap the  vessel  and  torque the cap to  12  ft-lbs  (16 N-m) or
      according to the unit  manufacturer's  directions.  Weigh the vessels  to the
      nearest 0.001 g.    Place the vessels  in the microwave carousel.

            CAUTION: Toxic  nitrogen  oxide  fumes  may be evolved, therefore all
            work must be performed in a properly operating ventilation system.
            The analyst  should also be aware of the  potential  for a  vigorous
            reaction.    If  a  vigorous reaction  occurs,  allow  to  cool before
            capping the  vessel.

            CAUTION:  When  digesting samples  containing  volatile  or easily
            oxidized organic compounds,  initially weigh no more  than 0.10 g and
            observe  the  reaction  before  capping the  vessel.   If  a  vigorous
            reaction occurs,  allow the  reaction to cease  before  capping the
            vessel.  If  no  appreciable  reaction  occurs,  a  sample  weight up to
            0.25 g can be used.
                                   3051 - 6                       Revision 0
                                                                  September 1994

-------
      CAUTION: All samples known or suspected of containing more than 5-
      10% organic material should be predigested in a hood for at least 15
      minutes.

      7.3.4   Properly place  the carousel  in the microwave unit  according
to the manufacturer's recommended specifications and, if used,  connect the
pressure  vessels  to  the  central overflow vessel  with PFA-fluorocarbon
tubes.  Any vessels containing  10 ml of nitric acid for analytical blank
purposes are counted  as sample  vessels.  When fewer than the  recommended
number  of  samples are to  be  digested, the remaining  vessels should be
filled  with 10  ml of  nitric  acid  to achieve  the full  complement of
vessels.   This  provides  an  energy  balance  since the  microwave  power
absorbed  is proportional  to the  total  mass  in  the  cavity  (Ref.  4).
Irradiate each group of sample vessels  for 10 minutes.   The temperature of
each  sample  should rise to 175 °C  in  less than 5.5 minutes and remain
between 170-180  °C for  the balance of  the 10 minute irradiation period.
The pressure  should  peak  at  less than 6 atm  for most  soil,  sludge,  and
sediment samples  (Ref. 5).  The  pressure will exceed these limits in the
case of high concentrations of  carbonate or organic compounds.  In these
cases the pressure will  be limited by the relief pressure of the  vessel to
7.5 ± 0.7 atm (110 ±  10 psi).   All  vessels should be sealed according to
the manufacturers recommended specifications.

            7.3.4.1  Newer microwave units are capable of higher  power (W)
      that permits digestion  of a larger  number  of  samples per batch.  If
      the analyst wishes to digest more samples at a time, the  analyst may
      use different  values of power  as long as they result  in  the same
      time  and  temperature  conditions defined  in  7.3.4.  That is,  any
      sequence of  power that  brings  the  samples to 175°C in  5.5 minutes
      and permits  a  slow rise  to  175  -  180°C during  the  remaining  4.5
      minutes (Ref. 5).

      Issues  of  safety,   structural   integrity  (both  temperature  and
      pressure limitations),  heat loss, chemical compatibility, microwave
      absorption  of  vessel   material,  and  energy  transport  will  be
      considerations made in  choosing  alternative vessels.  If all of the
      considerations  are met  and the appropriate power settings provided
      to reproduce the reaction conditions defined in  7.3.4,  then these
      alternative vessels may be used  (Ref. 1,2).

      7.3.5  At  the  end of the  microwave program,  allow the vessels to
cool for a minimum of 5 minutes before removing them from the microwave
unit.  When the vessels have cooled to  room temperature, weigh and record
the weight of each vessel  assembly.  If the weight of acid plus    sample
has decreased by more than 10 percent from the original     weight,
discard the sample.  Determine the reason for the weight loss.  These are
typically attributed to loss  of vessel  seal integrity, use of a digestion
time  longer  than 10  minutes,  too  large  a  sample, or  improper heating
conditions.  Once the source  of the loss  has  been     corrected, prepare
a new sample or set of samples for  digestion  beginning  at 7.3.1.


                             3051 -  7                       Revision  0
                                                            September  1994

-------
            7.3.6  Complete the preparation of the sample by carefully uncapping
      and venting each vessel in a fume hood.  Transfer the sample to an acid-
      cleaned  bottle.  If the digested sample contains particulates which may
      clog  nebulizers or  interfere with  injection  of the  sample  into  the
      instrument, the sample may be centrifuged, allowed to settle, or filtered.

                  7.3.6.1   Centrifugation:   Centrifugation  at 2,000-3,000 rpm
            for 10 minutes is usually sufficient to clear the supernatant.

                  7.3.6.2   Settling:   Allow  the  sample to  stand  until  the
            supernatant is  clear.   Allowing  a sample  to stand  overnight will
            usually accomplish this.  If it does not, centrifuge or filter the
            sample.

                  7.3.6.3    Filtering:    The  filtering  apparatus  must  be
            thoroughly cleaned and prerinsed with dilute (approximately 10% V/V)
            nitric acid.  Filter the sample  through   qualitative filter paper
            into a second acid-cleaned  container.

            7.3.7  Dilute  the digest to a known volume ensuring that the samples
      and standards are matrix matched.  The  digest  is now  ready for analysis
      for elements of interest using the appropriate SVI-846 method.

      7.4  Calculations:  The concentrations determined are to be  reported  on the
basis of the actual weight of the original  sample.

8.0  QUALITY CONTROL

      8.1   All  quality control  data  must  be  maintained  and   available  for
reference or inspection for  a period of three years.  This method is restricted
to  use  by,  or  under supervision  of,  experienced  analysts.   Refer to  the
appropriate section of Chapter One for  additional quality control guidance.

      8.2  Duplicate samples  should be processed on a routine basis.  A duplicate
sample is a sample brought through the  whole sample preparation  and analytical
process.  A duplicate sample should be  processed with each  analytical batch or
every 20 samples,  whichever  is the greater number. A  duplicate sample should be
prepared for each matrix type (i.e., soil,  sludge, etc.).

      8.3  Spiked  samples or standard reference materials should be included with
each group of  samples  processed or  every 20  samples,  whichever  is  the greater
number.   A spiked sample should also be  included whenever a new sample matrix is
being analyzed.

9.0  METHOD PERFORMANCE

      9.1   Precision:   Precision data for  Method 3051, as determined  by the
statistical examination of inter!aboratory test results,  is located in Tables 1
and 2.


                                   3051 - 8                       Revision 0
                                                                  September 1994

-------
      9.2  Repeatability:  If successive results are obtained by the same analyst
with the same  apparatus  under  constant operating conditions on identical test
material, then the difference between these successive results will  not,  with 95%
probability, exceed the repeatability value.   For example,  in  the  case  of lead,
an average of only 1 case in 20 would exceed

                                    0.206 x

in the long run, where x is one result in /xj/g (Ref.  6).

      9.3  Reproducibility: If two successive measurements are made  independently
by each of two  different analysts working in different laboratories on identical
test material, then the difference between the average result for  each analyst
will not, with 95% probability, exceed the reproducibility  value.   For example,
in the case of lead, an average of only  1 case in 20 would exceed

                                    0.303 x

in the long run, where x is the average  of two successive  measurements in /yg/g
(Ref. 2).

      As can be seen in Table  1, repeatability and reproducibility differ between
elements, and usually depend  on that element's  concentration.  Table 2 provides
an  example of  how  users  of  the method  can determine  expected  values  for
repeatability and reproducibility; nominal values of lead have been  used  for this
model (Ref. 6).

      9.4  Bias:   In  the case of SRM 1085 - Wear Metals  in Oil, the bias of this
test method is different  for each element.  An  estimate of  bias, as  shown in
Table 3, is:

                    Bias = Amount found  - Amount expected.

      However,  the  bias  estimate  inherits  both  the  uncertainty  in  the
measurements made using Method 3051 and the uncertainty on the certificate, so
whether the bias  is  real  or only due  to measurement error must  also  be con-
sidered.   The  concentrations  found  for Al, Cr,  and Cu  using  Method 3051 fall
within their certified ranges on SRM 1085, and 95% confidence intervals for Fe
and Ni overlap with their respective certified ranges;  therefore,   the observed
biases for these elements are  probably due  to  chance  and  should be considered
insignificant.  Biases  should not be estimated at  all for Ag  and Pb  because these
elements were  not certified.  Therefore, the only two elements considered in this
table for which the bias estimates are significant are Mg and Mo.

10.0  REFERENCES

1.    Test Methods for Evaluating Solid  Waste,  Physical/Chemical  Methods,  3rd
      ed;  U.S.  Environmental  Protection Agency, Office   of  Solid Waste  and
      Emergency Response.  U.S.  Government Printing Office:   Washington,  DC,
      1986; SW-846.


                                   3051 - 9                       Revision 0
                                                                  September 1994

-------
2.    Kingston, H. M.  and  L. B. Jassie, "Safety Guidelines  for Microwave Systems
      in  the  Analytical   Laboratory".    In  Introduction  to  Microwave  Acid
      Decomposition:  Theory  and  Practice;  Kingston,  H.  M.  and Jassie,  L. B.,
      eds.; ACS Professional Reference Book Series; American Chemical Society:
      Washington, DC, 1988.

3.    1985 Annual Book of ASTM  Standards,  Vol. 11.01; "Standard Specification
      for Reagent Water ; ASTM,  Philadelphia,  PA, 1985, D1193-77.

4.    Introduction  to  Microwave  Sample  Preparation:  Theory  and  Practice,
      Kingston, H. M. and Jassie, L. B., Eds.; ACS Professional Reference Book
      Series; American Chemical  Society:  Washington, DC, 1988.

5.    Kingston,  H.  M.    EPA  IAG #DWI-393254-01-0  January  1-March  31,  1988,
      quarterly Report.

6.    Binstock, D. A., Yeager, W. M.,  Grohse,  P.  M.  and Gaskill, A.  Validation
      of a Method for Determining Elements  in Solid Waste by Microwave Diges-
      tion,  Research  Triangle  Institute Technical  Report  Draft,  RTI  Project
      Number 321U-3579-24,  November,  1989,  prepared  for the Office  of Solid
      Waste, U.S. Environmental  Protection Agency, Washington, DC 20460.

7.    Kingston,  H.  M.,  Walter,  P.  J.,   "Comparison  of  Microwave  Versus
      Conventional Dissolution  for  Environmental Applications",  Spectroscopy,
      vol. 7 No. 9,20-27,1992.
                                   3051  -  10                      Revision 0
                                                                  September 1994

-------
                                  TABLE 1.
         EQUATIONS RELATING REPEATABILITY AND REPRODUCIBILITY TO MEAN
     CONCENTRATION OF DUPLICATE  DETERMINATION  WITH  95 PERCENT CONFIDENCE
            Element          Repeatability          Reproducibility
Ag
Al
B
Ba
Be
Ca
Cd
Co
Cr
Cu
Fe
Mg
Mn
Mo
Ni
Pb
Sr
V
Zn
0.195X8
0.232X
12. 9b
0.238X
0.082b
0.356X
0.385X
0.291X
0.187X
0.212X
0.257X
0.238X
1.96X1/2°
0.701X
0.212X
0.206X
0.283X
1.03X1/2
3.82X1/2
0.314X
0.444X
22. 6b
0.421X
0.082b
1.27X
0.571X
0.529X
0.195X
0.322X
0.348X
0.399X
4.02X1/2
0.857X
0.390X
0.303X
0.368X
2.23X1/2
7.69X1/2
"Log transformed variable based on one-way analysis of variance.
""Repeatability and reproducibility were independent of concentration.
°Square root transformed variable based on one-way  analysis  of variance.
                                  3051 - 11                       Revision 0
                                                                  September 1994

-------
                                    TABLE  2.
                  REPEATABILITY AND REPRODUCIBILITY FOR LEAD
                                BY  METHOD 3051
         Average Value       Repeatability      Reproducibilitv
50
100
200
300
400
500
10.3
20.6
41.2
61.8
82.4
103
15.2
30.3
60.6
90.9
121
152
All results are in mg/Kg
                                   3051 - 12                       Revision 0
                                                                   September 1994

-------
                                   TABLE 3.
           RECOVERY AND BIAS DATA FOR SRM  1085  - WEAR METALS  IN OIL
     Element
 Amount
 Expected
(Certified
  Range)
 Amount
  Found*
(95% Conf
Interval)
                                        Absolute
                                          Bias
         Relative
          Bias
        (Percent)
         Significant
         (due to more
         than chance)
       Ag
       Al
       Cr
       Cu
       Fe
       Mg
       Mo
       Ni
       Pb
  (291)**
  296+4
  298±5
  295±10
  300±4
  297±3
  292+11
  303+7
  (305)**
All values in mg/Kg
 234±16
 295+12
 293±10
 289±9
 311+14
 270±11
 238+11
 293+9
 279±8
-1
-5
-6
+11
-27
-54
-10
 0
-2
-2
+4
-9
-18
-3
No
No
No
No
Yes
Yes
No
 *Results taken from table 4-7, Ref. 2.

 **Value not certified,  so should not be used in bias detection and estimation.
                                   3051  -  13
                                                   Revision 0
                                                   September 1994

-------
                                 METHOD 3051
(MICROWAVE ASSISTED ACID DIGESTION OF SEDIMENTS,  SLUDGES, SOILS,  AND OILS)
                           Hn
                           weight
                         decreased >
                          10% from
                          original?
                                   3051 - 14
Revision  0
Septenter 1994

-------
                                 METHOD 3540

                             SOXHLET EXTRACTION
1.0  SCOPE AND APPLICATION

     1.1  Method 3540 is  a  procedure  for  extracting  nonvolatile and semi-
volatile organic compounds from  solids  such  as  soils,  sludges,  and wastes.
The Soxhlet extraction process ensures  intimate  contact of the sample matrix
with the extraction solvent.

     1.2  This method is  applicable  to  the  isolation  and concentration of
water-insoluble and  slightly  water-soluble  organics  in  preparation  for a
variety of chromatographic procedures.


2.0  SUMMARY OF METHOD

     2.1  The solid sample is mixed  with  anhydrous sodium sulfate, placed in
an extraction thimble or between two  plugs of glass wool, and extracted using
an appropriate solvent  in a  Soxhlet  extractor.    The extract is then dried,
concentrated, and, as necessary, exchanged  into a solvent compatible with the
cleanup or determinative step being employed.


3.0  INTERFERENCES

     3.1  Refer to Method 3500.
4.0  APPARATUS AND MATERIALS

     4.1  Soxhlet extractor;  40-mm  I.D., with 500-mL round-bottom flask.

     4.2  Drying column;   20-mm   I.D.   Pyrex chromatographic column with Pyrex
glass wool at bottom and  a Teflon  stopcock.
     NOTE:   Fritted glass  discs   are   difficult to decontaminate after highly
     contaminated extracts have been  passed  through.   Columns without frits
     may  be  purchased.  Use  a  small   pad  of  Pyrex glass wool to retain the
     adsorbent.  Prewash  the  glass wool pad  with 50 mL of acetone followed by
     50 mL of elution  solvent prior  to  packing the column with adsorbent.

     4.3  Kuderna-Danish  (K-D) apparatus;

          4.3.1  Concentrator tube:  10-mL, graduated (Kontes K-570050-1025 or
     equivalent).  Ground-glass   stopper  is  used  to  prevent evaporation of
     extracts.

          4.3.2  Evaporation   flask:      500-mL   (Kontes   K-570001-500  or
     equivalent).  Attach to  concentrator tube with springs.
                                   3540 -  1
                                                         Revision      0
                                                         Date  September 1986

-------
          4.3.3  Snyder column:     Three-ball   macro  (Kontes  K-503000-0121  or
     equivalent).

          4.3.4  Snyder  column:     Two-ball   micro  (Kontes   K-569001-0219  or
     equivalent).

     4.4  Boiling chips:  Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

     4.5  Water  bath;    Heated,  with  concentric  ring  cover,   capable  of
temperature control[+5*C).  The  bath should  be used in a hood.

     4.6  Vials;   Glass, 2-mL capacity, with  Teflon-lined screw cap.

     4.7  Glass or paper thimble or glass wool;  Contaminant  free.

     4.8  Heating mantle;  Rheostat controlled.

     4.9  Syringe;  5-mL.

     4.10  Apparatus  for determining percent moisture;

          4.10.1  Oven:  Drying.

          4.10.2  Desiccator.

          4.10.3   Crucibles:  Porcelain.

     4.11  Apparatus  for grinding:   If the sample will not pass through a 1-mm
 standard  sieve or cannot   be  extruded  through  a   1-mm  opening,  it should be
 processed into a homogeneous  sample   that  meets   these  requirements.  Fisher
 Mortar Model  155 Grinder,  Fisher Scientific Co., Catalogue Number  8-323,  or an
 equivalent brand  and  model,  is  recommended   for  sample   processing.  This
 grinder should handle  most   solid   samples,   except  gummy,  fibrous, or oily
 materials.


 5.0 REAGENTS

     5.1   Reagent water;    Reagent   water is   defined  as  water  in which an
 interferent  is not observed  at the   method detection limit of the  compounds of
 interest.

     5.2   Sodium sulfate;   (ACS) Granular anhydrous (purified  by  washing with
 methylene chloride followed  by heating at 400*C for 4 hr in a shallow tray).

     5.3   Extraction  solvents;

           5.3.1   Soil/sediment  and  aqueous   sludge  samples shall  be extracted
     using  either of  the following  solvent systems.
                                   3540 - 2
                                                          Revision       0
                                                          Date   September 1986

-------
               5.3.1.1  Toluene/Methanol:    10:1   (v/v),   pesticide  quality  or
          equivalent.

               5.3.1.2  Acetone/Hexane:     1:1   (v/v),   pesticide  quality   or
          equivalent.

          5.3.2  Other samples shall  be  extracted using  the following:

               5.3.2.1  Methylene chloride:   pesticide  quality or  equivalent.

     5.4  Exchange solvents;    Hexane,   2-propanol,  cyclohexane,  acetonitrile
(pesticide quality or equivalent).


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See the introductory  material   to  this  chapter, Organic Analytes,
Section 4.1.


7.0  PROCEDURE

     7.1  Sample handling;

          7.1.1  Sediment/soil samples:   Decant and discard any water layer on
     a sediment sample.  Mix sample thoroughly, especially composited samples.
     Discard any foreign objects such as sticks,  leaves, and rocks.

          7.1.2  Waste samples:   Samples  consisting  of  multiphases must be
     prepared by the phase  separation method in Chapter Two before extraction.
     This procedure is for  solids only.

          7.1.3  Dry waste  samples amenable  to  grinding:  Grind or otherwise
     subdivide the waste so that it either  passes through a 1-mm sieve or can
     be extruded through a  1-mm  hole.    Introduce sufficient sample into the
     grinding apparatus to  yield at  least 10 g after grinding.

     7.2  Determination of  percent moisture;   In certain cases, sample results
are desired  based on  a dry-weight basis.  When such data is desired, a portion
of sample for moisture determination  should be weighed out  at the same time as
the portion  used  for  analytical  determination.

          7.2.1   Immediately  after weighing  the  sample for extraction, weigh
     5-10 g  of  the   sample  into  a  tared  crucible.   Determine  the percent
     moisture by  drying overnight at  105*C.    Allow  to  cool in a desiccator
     before  weighing:
                                   3540 - 3
                                                         Revision      0
                                                         Date  September 1986

-------
     7.3  Blend 10 g of the solid sample with 10 g of anhydrous  sodium sulfate
and place in an extraction thimble.   The extraction thimble must  drain freely
for the duration of the extraction period.   A glass wool  plug above and below
the sample in  the  Soxhlet  extractor  is  an  acceptable alternative for the
thimble.  Add 1.0  ml  of  the  surrogate  standard  spiking solution onto the
sample (See Method  3500  for  details  on  the  surrogate standard and matrix
spiking solutions.)  For  the  sample  in  each  analytical  batch  selected for
spiking, add 1.0 ml  of  the  matrix  spiking standard.   For base/neutral-acid
analysis, the amount  added  of  the  surrogates  and matrix spiking compounds
should result in  a  final  concentration  of  100  ng/uL of each  base/neutral
analyte and 200 ng/uL  of  each  acid  analyte  in  the  extract  to be analyzed
(assuming a 1 uL injection).  If Method 3640, Gel-permeation cleanup, is to be
used, add twice the volume  of  surrogates  and matrix spiking compounds since
half the extract is lost due to loading of the GPC column.

     7.4  Place 300 ml of the  extraction  solvent (Section 5.3) into a 500-mL
round-bottom flask containing one  or  two  clean  boiling  chips.  Attach the
flask to the extractor and extract the sample for 16-24 hr.

     7.5  Allow the extract to cool after the extraction is complete.

     7.6  Assemble a Kuderna-Danish   (K-D)  concentrator  by attaching a 10-mL
concentrator tube to a 500-mL evaporation flask.

     7.7  Dry the extract by  passing  it  through  a drying column containing
about 10 cm of anhydrous  sodium sulfate.    Collect the dried extract in a K-D
concentrator.  Wash the extractor  flask and sodium sulfate column with
100-125 ml of extraction  solvent to complete the quantitative transfer.

     7.8  Add one or two  clean boiling chips  to the flask and attach a three-
ball Snyder column.     Prewet  the Snyder  column  by  adding  about   1 ml of
methylene chloride  to  the top of the  column.  Place the K-D  apparatus on a hot
water bath  (15-20'C  above   the  boiling  point  of  the  solvent)  so that the
concentrator tube  is partially immersed  in   the  hot water and the  entire lower
rounded  surface  of  the flask is  bathed   with   hot  vapor.  Adjust  the vertical
position of the  apparatus and the  water  temperature, as  required,  to complete
the  concentration  in  10-20  min.  At the  proper  rate of distillation,  the balls
of the  column will  actively chatter,   but  the   chambers will not  flood.  When
the  apparent volume of liquid reaches 1   ml,  remove the  K-D  apparatus  from the
water  bath  and  allow  it to  drain  and  cool for at least 10 min.

      7.9  If a   solvent  exchange   is  required  (as  indicated   in  Table 1),
momentarily  remove the Snyder column, add 50 ml of the  exchange  solvent  and a
new  boiling  chip,  and  re-attach  the  Snyder  column.  Concentrate the extract as
described  in Paragraph 7.6,   raising   the  temperature   of   the water bath, if
necessary,  to  maintain proper distillation.

      7.10   Remove the  Snyder column  and  rinse   the flask and its  lower joints
into the concentrator  tube  with   1-2  ml   of   methylene chloride or exchange
solvent.   If  sulfur  crystals   are   a  problem,  proceed  to  Method 3660 for
cleanup.   The   extract  may  be   further concentrated   by  using  the technique
outlined in  Paragraph  7.9 or adjusted to 10.0 ml with the solvent  last  used.


                                   3540 - 4
                                                          Revision       0
                                                          Date   September 1986

-------
   TABLE 1.   SPECIFIC  EXTRACTION  CONDITIONS  FOR VARIOUS DETERMINATIVE METHODS



Determinative
method
8040a
8060
8080
8090
8100
8120
8140
S250Y
8270a,C
8310



Extraction
PH
as received
as received
as received
as received
as received
as received
as received
as received
as received
as received
Exchange
solvent
required
for
analysis
2-propanol
hexane
hexane
hexane
none
hexane
hexane
none
none
acetonitrile
Exchange
solvent
required
for
cleanup
hexane
hexane
hexane
hexane
cyclohexane
hexane
hexane
_
-
~"
Volune
of extract
required
for
cleanup (mL)
1.0
2.0
10.0
2.0
2.0
2.0
10.0
_
-
"
Final
extract
volune
for
analysis (mL)
1.0, 10. Ob
10.0
10.0
1.0
1.0
1.0
10.0
1.0
1.0
1.0
 10 obtain separate acid and base/neutral extracts, Method 3650  should be performed following
concentration of the extract to 10.0 mL.

 Phenols may be analyzed, by Method 8040, using a 1.0 mL 2-propanol extract by GC/FID.  Method 8040
also contains an optional derivatization procedure for phenols which results in a 10 mL hexane
extract to be analyzed by GC/ECD.

The specificity of OC/MS may make cleanup of  the extracts unnecessary.  Refer to Method 3600 for
guidance on the cleanup procedures available if required.
                                           3540  - 5
                                                                      Revision       0	
                                                                      Date  September  1986

-------
     7.11  If further concentration Is indicated  in  Table 1,  add another one
or two clean boiling  chips  to  the  concentrator  tube and attach a two-ball
micro Snyder column.  Prewet the column by adding 0.5 ml of methylene chloride
or exchange solvent to the top of  the  column.   Place the K-D apparatus in a
hot water bath so that the concentrator  tube is partially immersed in the hot
water.    Adjust  the  vertical  position  of  the  apparatus  and  the  water
temperature, as required, to complete the  concentration  in 5-10 min.  At the
proper rate of distillation the balls of the column will actively chatter, but
the chambers will not flood.  When the apparent volume of liquid reaches
0.5 ml, remove the K-D apparatus from the water bath and allow it to drain and
cool for at least 10 min.   Remove  the  Snyder column and rinse the flask and
its lower joints into the concentrator  tube  with  0.2 mL of solvent.  Adjust
the final volume to 1.0-2.0 ml, as Indicated 1n Table 1, with solvent.

     7.12  The extracts obtained may now be analyzed for analyte content using
a variety of  organic  techniques  (see  Section  4.3  of  this  chapter).  If
analysis of  the  extract  will  not  be  performed  immediately,  stopper the
concentrator tube and  store  refrigerated.    If  the  extract will be stored
longer than 2 days, it should be transferred to a Teflon-sealed screw-cap vial
and labeled appropriately.


8.0 |QUALITY CONTROL

     8.1  Any reagent blanks or  matrix  spike  samples should be subjected to
exactly the same analytical procedures as those used on actual samples.

     8.2  Refer to  Chapter  One  for  specific  quality control procedures and
Method 3500 for extraction and sample preparation procedures.


9.0  METHOD PERFORMANCE

     9.1  Refer to  the determinative methods for performance data.


10.0   REFERENCES

1.  U.S.  EPA  40  CFR Part 136,  "Guidelines  Establishing  Test  Procedures  for the
Analysis  of Pollutants Under  the Clean Water Act;  Final  Rule and  Interim Final
Rule  and  Proposed  Rule," October 26,  1984.
                                   3540 - 6
                                                          Revision
                                                          Date  September 1986

-------
                                         METHOD  3540

                                     SOXHLET  EXTRACTION
 7. 1
       Use
   appropriate
•ample handling
   technique
                           7.6
 Assemble K-O
 concentrator
 7.2
    Determine
     percent
    moisture
 7.3
                           7. 7
                                                                              4.6
Dry and collect
extract In K-O
  concentrator
        Add
    appropriate
  surrogate  and
 matrix spiking
    standards
 7.4
                           4.6
   Concentrate
  using Snyder
column and K-D
   apparatus
        Place
      methylene
      chlor Jde:
     acetone  In
 flask;  extract
  for 16-24 hrs
                           Reconcentrate
                           using Snyder
                          column and K-O
                            apparatus
Analyze using
   organic
 techniques
                                     3540 -  7
                                                               Revision       p
                                                               Date  September  1986

-------
                                 METHOD 3550

                            SONICATION EXTRACTION
1.0  SCOPE AND APPLICATION

     1.1  Method 3550 Is  a  procedure  for  extracting  nonvolatile and semi-
volatlle organic compounds from  sol Ids  such  as  soils, sludges,  and wastes.
The sonication process ensures intimate contact  of the sample matrix with the
extraction solvent.

     1.2  The method is  divided  into  two  sections,  based  on the expected
concentration of  organics  in  the  sample.    The  low  concentration method
(Individual organic components of <[20 mg/kg)  uses  a larger sample size and a
more rigorous extraction procedure  (lower concentrations are more difficult to
extract).  The high concentration method (individual organic components of
>20 mg/kg) is much simpler and therefore faster.

     1.3  It is highly recommended that  the  extracts  be cleaned up prior to
analysis.  See Cleanup, Section 4.2.2 of Chapter Four, for applicable methods.


2.0  SUMMARY OF METHOD

     2.1  Low concentration method;   A  30-g  sample  1s mixed with anhydrous
sodium sulfate to form a free-flowing powder.  This is solvent extracted three
times using sonication.  The  extract  is  separated from the sample by vacuum
filtration or  centrifugation.    The  extract  is  ready  for  cleanup and/or
analysis following concentration.

     2.2  High concentration method;   A  2-g  sample  1s mixed with anhydrous
sodium sulfate to form a free-flowing  powder.  This is solvent extracted once
using sonication.  A  portion  of  the  extract  is removed for cleanup and/or
analysis.


3.0  INTERFERENCES

     3.1  Refer to Method 3500.


4.0  APPARATUS AND MATERIALS

     4.1  Apparatus for grinding;   If the  sample will not pass through a 1-mm
standard sieve or cannot  be  extruded  through  a  1-mm opening, it should be
processed into a homogeneous  sample  that  meets  these requirements.  Fisher
Mortar Model 155 Grinder, Fisher Scientific Co., Catalogue Number 8-323, or an
equivalent brand  and  model,  is   recommended  for  sample  processing.  This
grinder should handle  most  solid  samples,  except  gummy,  fibrous, or oily
materials.
                                  3550 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
     4.2  Sonication;  A  horn-type  sonicator  equipped  with  a titanium tip
should be used.  The following sonicator, or an equivalent brand and model, is
recommended:

          Ultrasonic cell disrupter:  Heat  Systems - Ultrasonics, Inc., Model
          W-385 (475 watt) sonicator  or  equivalent  (Power wattage must be a
          minimum of 375  with  pulsing  capability  and  No.  200 1/2" Tapped
          Disrupter Horn) plus No. 207 3/4" Tapped Disrupter Horn, and No. 419
          1/8" Standard Tapered microtip probe.

     4.3  Sonabox;    Recommended   with   above   disrupters  for  decreasing
          cavitation sound (Heat Systems  -  Ultrasonics,  Inc., Model 432B or
          equivalent).

     4.4  Apparatus  for determining percent moisture:

          4.4.1  Oven:  Drying.

          4.4.2  Desiccator.

          4.4.3  Crucibles:   Porcelain.

     4.5  Pasteur glass  pipets;   Disposable,  1-mL.

     4.6  Beakers:   400-mL.

     4.7  Vacuum filtration  apparatus:

          4.7.1  Buchner funnel.

          4.7.2  Filter paper:  Whatman  No.  41 or equivalent.

      4.8  Kuderna-Danish (K-D)  apparatus;

          4.8.1  Concentrator tube:   10-mL  graduated  (Kontes  K-570050-1025 or
      equivalent).

          4.8.2   Evaporator    flask:       500-mL   (Kontes   K-570001-0500  or
      equivalent).

          4.8.3  Snyder column:     Three-ball   macro  (Kontes  K-503000-0121 or
      equivalent).

          4.8.4   Snyder  column:     Two-ball   micro  (Kontes  K-569001-0219 or
      equivalent).

      4.9  Boiling  chips;  Solvent extracted,  approximately 10/40 mesh (silicon
 carbide or  equivalent).

      4.10  Water  bath:    Heated,  with  concentric  ring  cover,   capable of
 temperature control (15*C).   The bath should be used in a hood.
                                   3550 - 2
                                                          Revision
                                                          Date  September 1986

-------
     4.11  Balance:  Top-loading,  capable of accurately weighing 0.01 g.

     4.12  Vials and caps;  2-mL for GC auto-sampler.

     4.13  Glass scintillation vials:    At  least  20-mL,   with screw-cap and
Teflon or aluminum foil liner.

     4.14  Spatula:  Stainless steel or Teflon.

     4.15  Drying column;  20-mm I.D.  Pyrex chromatographic column with Pyrex
glass wool at bottom and a Teflon stopcock.
     NOTE:  Fritted glass  discs  are  difficult to decontaminate after highly
     contaminated extracts have been  passed  through.   Columns without frits
     may be purchased.  Use  a  small  pad  of  Pyrex glass wool to retain the
     adsorbent.  Prewash the glass wool pad  with 50 ml of acetone followed by
     50 ml of elution solvent prior to packing the column with adsorbent.

     4.16  Syringe;  5-mL.


5.0  REAGENTS

     5.1  Sodium sulfate:  Anhydrous and reagent grade, heated at 400*C for
4 hr, cooled in a desiccator, and   stored  in a glass bottle.  Baker anhydrous
powder, catalog #73898, or equivalent.

     5.2  Extraction   solvents:     Methylene   chloride:acetone   (1:1,  v:v),
methylene chloride, hexane (pesticide quality or equivalent).

     5.3  Exchange  solvents:    Hexane,  2-propanol, cyclohexane, acetonitrile
 (pesticide quality  or  equivalent).


6.0  SAMPLE COLLECTION, PRESERVATION, AND  HANDLING

     6.1  See the  introductory  material   to  this  chapter, Organic Analytes,
Section 4.1.
7.0  PROCEDURE

     7.1  Sample handling;

          7.1.1  Sediment/soil samples:  Decant and discard any water layer on
     a sediment sample.  Mix sample thoroughly, especially composited samples.
     Discard any foreign objects such as sticks, leaves, and rocks.

          7.1.2  Waste samples:   Samples  consisting  of  multiphases must be
     prepared by the phase separation method in Chapter Two before extraction.
     This procedure is for solids only.
                                  3550 - 3
                                                         Revision      0
                                                         Date  September 1986

-------
          7.1.3  Dry waste samples amenable  to  grinding:   Grind  or otherwise
     subdivide the waste so that it either  passes  through  a 1-mrn  sieve  or  can
     be extruded through a 1-mm  hole.     Introduce sufficient sample into  the
     grinding apparatus to yield at least 10 g after grinding.

     7.2  Determination of percent moisture;  In certain cases,  sample results
are desired based on a dry-weight basis.   When such data is desired, a portion
of sample for moisture determination should be weighed out  at the  same time as
the portion used for analytical determination.

          7.2.1  Immediately after weighing  the  sample for extraction, weigh
     5-10 g of  the  sample  into  a  tared  crucible.   Determine the percent
     moisture by drying overnight at  105*C.    Allow  to cool in  a desiccator
     before weighing:
               q of sampl- qdry sample x 100 = % mo1sture
     7.3  Determination of pH (if required):    Transfer  50  g of sample to a
100-mL beaker.  Add 50 ml of  water  and  stir  for 1 hr.  Determine the pH of
sample with glass electrode and pH meter while stirring.  Discard this portion
of sample.

     7.4  Extraction method for samples expected to contain low concentrations
of organics and pesticides  «20 mg/kg);

          7.4.1  The following step should be  performed rapidly to avoid loss
     of the more volatile extractables.    Weigh  approximately 30 g of sample
     into a 400-mL beaker.    Record  the  weight  to  the nearest 0.1 g.  Non-
     porous or wet samples   (gummy  or  clay  type)  that  do not have a free-
     flowing  sandy texture  must be mixed with 60 g of anhydrous sodium sulfate
     using a  spatula.   The  sample  should  be free-flowing at this point.  Add
     1 ml of  surrogate  standards  to  all  samples,   spikes, and blanks  (see
     Method 3500 for details  on the surrogate standard  solution and the matrix
     spike solution).   For  the  sample  in  each analytical batch selected for
     spiking, add  1.0  ml  of  the  matrix  spiking standard.  For base/neutral-
     acid analysis, the  amount  added  of  the  surrogates and matrix spiking
     compounds  should  result  in  a  final   concentration  of 100 ng/uL of each
     base/neutral  analyte and 200 ng/uL of  each acid analyte in the extract to
     be analyzed  (assuming  a  1 uL  injection).  If Method 3640, Gel -permeation
     cleanup, is to be used,  add  twice  the  volume of surrogates and matrix
     spiking  compounds since  half of the extract is  lost due to loading of the
     GPC  column.   Immediately add 100 ml of 1:1 methylene chloride:acetone.

          7.4.2  Place the  bottom  surface  of  the  tip  of  the #207 3/4 in.
     disrupter  horn about 1/2 in. below  the surface of the solvent, but above
     the  sediment  layer.

          7.4.3  Sonicate for 3 min, with   output  control  knob set at  10 and
     with mode  switch  on  Pulse and percent-duty cycle knob  set at 50%.  Do NOT
     use  microtip  probe.
                                   3550 - 4
                                                          Revision
                                                          Date   September  1986

-------
     7.4.4  Decant and  filter  extracts  through  Whatman  No.  41  filter
paper  using  vacuum  filtration  or  centrifuge  and  decant  extraction
solvent.

     7.4.5  Repeat the extraction two  or  more times with two additional
100-mL portions of solvent.  Decant off the extraction solvent after each
sonication.  On the  final  sonication,  pour  the entire sample into the
Buchner funnel and rinse with extraction solvent.

     7.4.6  Assemble a Kuderna-Danish  (K-D)  concentrator by attaching a
10-mL concentrator tube to a 500-mL evaporative flask.

     7.4.7  Dry  the  extract  by  passing  it  through  a  drying  column
containing about 10 cm of  anhydrous  sodium  sulfate.  Collect the dried
extract in a  K-D  concentrator.    Wash  the  extractor flask and  sodium
sulfate column with  100-125  ml  of  extraction  solvent to complete the
quantitative transfer.

     7.4.8  Add one or two clean  boiling  chips to the evaporative flask
and attach a  three-ball  Snyder  column.    Prewet  the Snyder column by
adding about 1 ml methylene chloride to the top.  Place the K-D apparatus
on a hot water bath (80-90*C)  so that the concentrator tube is partially
immersed in the hot water  and  the  entire  lower rounded surface  of the
flask is bathed with  hot  vapor.    Adjust  the vertical position  of the
apparatus  and  the  water  temperature,  as  required,  to  complete the
concentration in 10-15 min.  At the proper rate of distillation the balls
of the column will actively chatter, but the chambers will not flood with
condensed  solvent.  When  the  apparent  volume  of  liquid reaches 1 ml,
remove the K-D apparatus  and allow it to drain and cool for at least
10 min.

     7.4.9 If a solvent  exchange is  required  (as indicated in Table 1),
momentarily remove the Snyder column,  add  50 ml of the exchange solvent
and a new  boiling chip, and re-attach the Snyder column.  Concentrate the
extract as described  in Paragraph  7.4.8,   raising the temperature of the
water bath, if necessary,  to maintain proper distillation.

     7.4.10   Remove the Snyder column and   rinse  the flask and its lower
joints  into the concentrator tube  with   1-2  mL of methylene chloride or
exchange  solvent.   If sulfur  crystals   are  a  problem, proceed to Method
3660 for  cleanup.   The  extract   may  be  further concentrated by using the
technique  outlined  in Paragraph  7.4.11   or  adjusted  to  10.0 ml with the
solvent last  used.

     7.4.11   Add a  clean  boiling chip   and  attach a two-ball micro-Snyder
column  to  the concentrator tube.     Prewet  the column by adding approxi-
mately  0.5 ml of methylene  chloride  or  exchange  solvent  through the top.
Place the  apparatus in  the  hot   water  bath.  Adjust the vertical position
and the water temperature,  as   required,  to complete  the  concentration in
5-10 min.  At the proper   rate   of   distillation, the balls of the column
will actively chatter,  but  the  chambers  will not  flood.  When the liquid
                              3550 -  5
                                                    Revision      0
                                                    Date  September 1986

-------
   TABLE 1.   SPECIFIC  EXTRACTION  CONDITIONS  FOR VARIOUS  DETERMINATIVE METHODS



Determinative
method
80403
8060
8080
8090
8100
8120
8140
8250a,C
8270a,C
8310



Extraction
pH
as received
as received
as received
as received
as received
as received
as received
as received
as received
as received
Exchange
solvent
required
for
analysis
2-propanol
hexane
hexane
hexane
none
hexane
hexane
none
none
acetonitrile
Exchange
solvent
required
for
cleanup
hexane
hexane
hexane
hexane
cyclohexane
hexane
hexane
_
-
"
Volune
of extract
required
for
cleanup (mL)
1.0
2.0
10.0
2.0
2.0
2.0
10.0
_
-
"
Final
extract
volume
for
analysis (mL)
1.0, 10. Ob
10.0
10.0
1.0
1.0
1.0
10.0
1.0
1.0
1.0
 TO obtain separate acid and base/neutral extracts, Method 3650 should be performed following
concentration of the extract to 10.0 mL.

 Phenols may be analyzed, by Method 8040, using a 1.0 mL 2-propanol extract by GC/FID.  Method 8040
also contains an optional derivatization procedure  for phenols which  results in a 10 mL hexane
extract to be analyzed by GC/ECD.

The specificity of GC/MS may make cleanup of the extracts unnecessary.  Refer to Method 3600 for
guidance on the cleanup procedures available if required.
                                           3550 - 6
                                                                       Revision       0	
                                                                       Date   September  1986

-------
     reaches an apparent volume of  approximately  0.5  ml,  remove  the  apparatus
     from the water bath and  allow  to   drain   and  cool  for  at  least  10  min.
     Remove the  micro-Snyder  column and  rinse  its  lower   joint   into the
     concentrator tube  with  approximately  0.2  ml  of  appropriate solvent.
     Adjust the final  volume to  the  volume  required  for cleanup or  for the
     determinative method (see Table 1).

          7.4.12  Transfer the concentrated extract  to a clean screw-cap vial.
     Seal the vial with a  Teflon-lined   lid  and  mark the level on  the vial.
     Label with the sample number and  fraction  and  store in the dark at 4*C
     until ready for analysis or cleanup.

     7.5  Extraction method  for  samples  expected   to  contain   high  concen-
trations of organics (>20 mg/kg):

          7.5.1  Transfer approximately  2 g (record  weight to  the nearest
     0.1 g) of sample to a  20-mL  vial.     Wipe  the  mouth of the vial with a
     tissue to remove any sample material.   Record  the exact  weight  of sample
     taken.  Cap the vial before proceeding  with  the  next sample to  avoid any
     cross contamination.

          7.5.2  Add 2 g of anhydrous  sodium  sulfate  to sample in  the 20-mL
     vial and mix well.

          7.5.3  Surrogate standards are  added  to   all  samples, spikes, and
     blanks  (see Method 3500  for  details  on the surrogate standard solution
     and on the matrix  spike  solution).    Add  2.0   mL of surrogate spiking
     solution to sample mixture.    For  the  sample  in each  analytical batch
     selected for spiking, add 2.0  mL  of  the  matrix spiking standard.  For
     base/neutral-acid analysis, the amount added  of the surrogates and matrix
     spiking compounds should result in  a  final concentration of 200 ng/uL of
     each base/neutral analyte  and  400  ng/uL  of   each  acid analyte in the
     extract to be analyzed (assuming a 1 uL injection).  If Method 3640, Gel-
     permeation cleanup, is to be used,  add twice  the  volume of surrogates and
     matrix spiking compounds since half the extract is lost due  to loading of
     the GPC column.

          7.5.4  Immediately add whatever  volume  of  solvent is necessary to
     bring the  final  volume  to  10.0  mL  considering  the  added volume of
     surrogates and  matrix  spikes.    Disrupt  the  sample  with the 1/8-in.
     tapered microtip ultrasonic probe for  2  min at output control  setting 5
     and with mode switch on pulse and  percent duty cycle of 50%.  Extraction
     solvents are:

          1.  Nonpolar compounds, i.e.,  organochlorine pesticides and
              PCBs:  hexane.

          2.  Extractable priority pollutants:  methylene chloride.

          7.5.5   Loosely pack disposable  Pasteur pipets  with 2- to 3-cm  Pyrex
     glass-wool plugs.   Filter the extract  through the glass wool and collect
                                  3550 - 7
                                                         Revision
                                                         Date  September 1986

-------
     5.0 ml in  a  concentrator  tube  if  further  concentration  is  required.
     Follow Paragraphs  7.4.6  through  7.4.12  for  details  on concentration.
     Normally, the 5.0 ml extract is concentrated to 1.0 ml.

          7.5.6  The extract is ready  for  cleanup  or analysis,  depending on
     the extent of interfering co-extractives.


8.0  QUALITY CONTROL

     8.1  Any reagent blanks  or  matrix  spike  samples  should be subject to
exactly the same analytical procedures as those used on actual samples.

     8.2  Refer to Chapter  One  for  specific  quality control procedures and
Method 3500 for extraction and sample preparation procedures.


9.0  METHOD PERFORMANCE

     9.1  Refer to the determinative methods  for performance data.


10.0  REFERENCES

1.  U.S.  EPA  40 CFR  Part  136,  "Guidelines Establishing  Test  Procedures for the
Analysis  of Pollutants Under  the Clean Water  Act;  Final  Rule and  Interim Final
Rule and  Proposed  Rule,"  October 26,  1984.

2.  U.S.  EPA,  Interlaboratory Comparison Study: Methods  for  Volatile and Semi-
Volatile  Compounds,   Environmental   Monitoring  Systems  Laboratory, Office of
Research  and  Development,  Las Vegas,  NV, EPA  600/4-84-027, 1984.
                                   3550 - 8
                                                          Revision
                                                          Date  September 1986

-------
                                      ,  METHOD 3550

                                   SONICATION EXTRACTION
    Start
7. 1
      Prepare
 samples using
  approprlate
method tor the
 waste ma tr1x
7 .2
                         7 .5.3
                       Add anhydrous
                       sod lum sul1 ate
                         to sample
     Determine
   the percent
of moisture In
   the sample
7.3
                         7,5.3
                                                                             7.4.1
        Add
      surrogate
   standards to
   •11 samples.
   spikes,  and
      blanks
                               Add
                            surrogate
                            standards
                      to all samples.
                         spikes,  and
                             blanks
 Determine pH
  of sample
                                                                             7.4
Sonicate sample
  •t least 3
     times
                         7.5.4   Adjust
                                 vo lume;
                       disrupt sample
                        with tapered
                      mlcrotlp ultra-
                        sonic probe
o
                            16 further
                          concentration
                            required?
s


7.S.S
Fllte
gl<

r through
ss wool




7.4.8
Cone
extr
collec
conce
                                                   7.4.9
                                                     Add exchange
                                                       solvent:
                                                     concentrate
                                                       extract
                                                                           Is • solvent
                                                                             exchange
                                                                            required?
                                    3550 - 9
                                                               Revision       0
                                                               Date  September 1986

-------
                                    METHOD 3S5O

                               SONICATION EXTRACTION
                                    (Continued)
o
                                                                      Use Matnod  366O
                                                                        tor cleanup
                                                   Partner
                                                 concentrate
                                                »nd/or adjust
                                                  volume
                                3550 - 10
                                                          Revision      0
                                                          Date  September  1986

-------
                                 METHOD 3580

                               WASTE DILUTION
1.0  SCOPE AND APPLICATION

     1.1  This method describes  a  solvent  dilution  of  a non-aqueous waste
sample prior to cleanup and/or analysis.     It is designed for wastes that may
contain organic chemicals at a  level   greater  than 20,000 mg/kg and that are
soluble in the dilution solvent.

     1.2  It is recommended that an  aliquot  of the diluted sample be cleaned
up.  See the Cleanup section of this chapter for methods (Section 4.2.2).


2.0  SUMMARY OF METHOD

     2.1  One gram of sample is weighed into  a capped tube, and the sample is
diluted to 10.0 mL with an appropriate solvent.


3.0  INTERFERENCES

     3.1  Refer to Method 3500.
4.0  APPARATUS AND MATERIALS

     4.1  Glass scintillation vials;  At least 20-mL, with Teflon or aluminum-
foil-lined screw-cap.,

     4.2  Spatula:  Stainless steel or Teflon.

     4.3  Balance;  Capable of weighing 100 g to the nearest 0.01 g.

     4.4  Vials and caps;  2-mL for GC autosampler.

     4.5  Disposable pi pets;  Pasteur.

     4.6  Test tube rack.

     4.7  Pyrex glass wool.

     4.8  Volumetric flasks;  10-mL (optional).


5.0  REAGENTS

     5.1  Sodium sulfate:  (ACS)  Granular,  anhydrous (purified by heating at
400*C for 4 hr in a shallow tray).
                                  3580 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
     5.2  Solvents:   Methylene  chloride  and  hexane  (pesticide  quality or
equi valentJT


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See the introductory  material  to  this  chapter,  Organic Analytes,
Section 4.1-.


7.0  PROCEDURE

     7.1  Samples consisting of  multiphases  must  be  prepared  by the phase
separation method  (Chapter Two) before extraction.

     7.2  The sample dilution may  be  performed  in a 10-mL volumetric flask.
If disposable glassware  is  preferred,  the  20-mL  scintillation vial may be
calibrated for use.   Simply  pi pet  10.0  mL  of  extraction solvent into the
scintillation vial and mark the bottom of the meniscus.  Discard this solvent.

     7.3  Transfer approximately 1  g  of  each  phase  (record  weight to the
nearest 0.1 g) of  the  sample  to  separate  20-mL  vials or 10-mL volumetric
flasks.  Wipe the  mouth  of  the  vial  with  a  tissue  to remove any sample
material.  Cap the vial before  proceeding  with  the next sample to avoid any
cross-contamination.

     7.4  Add 2.0  mL surrogate  spiking  solution  to  all samples and blanks.
For the sample in  each analytical  batch  selected  for spiking, add 2.0 mL of
the matrix spiking standard.  For base/neutral-acid analysis, the amount added
of  the surrogates  and  matrix  spiking  compounds  should  result  in a final
concentration of 200 ng/uL of each  base/neutral analyte and 400 ng/uL of each
acid analyte  in the extract to  be  analyzed  (assuming a 1 uL  injection).  If
Method 3640,  Gel-permeation cleanup, is  to   be  used, add twice the volume of
surrogates and matrix  spiking compounds  since half the extract is lost due to
loading of the GPC column.     See  Method  3500  for details on the surrogate
standard  and  matrix spiking solutions.

     7.5   Immediately  dilute  to  10  mL with the  appropriate  solvent.   For
compounds  to  be analyzed by   GC/ECD, e.g.,  organochlorine pesticides and PCBs,
the dilution  solvent should be  hexane.   For base/neutral  and acid semivolatile
priority  pollutants, use methylene chloride.    If dilution  is to  be cleaned up
by  gel permeation  chromatography  (Method  3640),  use  methylene  chloride as  the
dilution  solvent  for all compounds.

     7.6   Add 2.0  g of anhydrous  sodium sulfate  to  the  sample.

     7.7   Cap and  shake the  sample for 2 min.

     7.8   Loosely  pack disposable Pasteur pipets with 2-3 cm glass  wool plugs.
 Filter the extract through the glass wool  and collect 5 mL  of the extract  in  a
 tube or  vial.
                                   3580 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
     7.9  The extract is  ready  for  cleanup  or  analysis,   depending on the
extent of interfering co-extractives.

8.0  QUALITY CONTROL
     8.1  Any reagent blanks and matrix  spike  samples should be subjected to
exactly the same analytical procedures as those used on actual samples.
     8.2  Refer to Chapter  One  for  specific  quality control procedures and
Method 3500 for extraction and sample preparation procedures.

9.0  METHOD PERFORMANCE
     9.1  Refer to the determinative methods for performance data.

10.0  REFERENCES
      10.1  None applicable.
                                  3580 - 3
                                                         Revision      0
                                                         Date  September 1986

-------
                                          METHOD  3580

                                        WASTE  DILUTION
   Does sample
contain more than
   one phase?
                                Use phase
                               separation
                                 method
                              (chapter 2)
    Transfer i
   Of each phase
    to separate
 vials or flasks
      I    Add
     —I surrogate
        spiking
     solutIon to
     all samples
     and blanks
 matrix spiking
   standard to
sample selected
   for spiking
                                                                               7.5
Dilute with
approprlate
  solvent
                                                                               7.6
                                                                              Add anhydrous
                                                                                 ammonium
                                                                                 sulfate
                                                                              7.7 I


                                                                              Cap and shake
                                                                               7.8
                                                                               Filter  through
                                                                                 glass  wool
                                      3580 - 4
                                                                Revision        0
                                                                Date  September  1986

-------
                                 METHOD 3500

                  ORGANIC EXTRACTION AND SAMPLE PREPARATION
1.0  SCOPE AND APPLICATION

     1.1  The  3500  Methods  are  procedures  for  quantitatively  extracting
nonvolatile and semi volatile organic  compounds  from various sample matrices.
Cleanup and/or analysis of  the  resultant  extracts  are described in Chapter
Four, Sections 4.2.2 and 4.3, respectively.

     1.2  Method 3580 describes a solvent  dilution technique that may be used
on non-aqueous nonvolatile and  semi volatile  organic samples prior to cleanup
and/or analysis.

     1.3  The 5000 Methods  are  procedures  for  preparing samples containing
volatile organic compounds for quantitative analysis.

     1.4  Refer to the specific method of interest for further details.


2.0  SUMMARY OF METHOD

     2.1  3500 Methods;  A  sample  of  a  known  volume  or weight is solvent
extracted.  The resultant extract is dried and then concentrated in a Kuderna-
Danish apparatus.  Other concentration  devices  or  techniques may be used in
place of the Kuderna-Danish  concentrator  if the quality control requirements
of the determinative methods are met (Method 8000, Section 8.0).

     2.2  5000 Methods;  Refer to the specific method of interest.


3.0  INTERFERENCES

     3.1  Samples requiring analysis  for  volatile  organic compounds, can be
contaminated by  diffusion  of  volatile  organics  (particularly chlorofluoro-
carbons and methylene  chloride)  through  the  sample container septum during
shipment and storage.  A field  blank  prepared from reagent water and carried
through sampling and subsequent storage and  handling  can serve as a check on
such contamination.

     3.2  Solvents, reagent, glassware,  and  other sample processing hardware
may yield  artifacts   and/or  interferences  to  sample  analysis.   All these
materials must  be  demonstrated  to  be   free  from  interferences  under the
conditions of the analysis by analyzing  method blanks.  Specific selection of
reagents and purification of solvents by distillation in all-glass systems may
be required.  Refer to Chapter  One  for  specific  guidance  on quality control
procedures.

     3.3   Interferences coextracted from   the  samples  will vary considerably
from source  to  source.  If analysis of an  extracted sample is prevented due to
interferences,  further cleanup of the sample  extract may be necessary.  Refer
to Method  3600  for  guidance on cleanup procedures.

                                  3500 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
     3.4  Phthalate esters contaminate many  types  of products commonly found
in  the  laboratory.    Plastics,  in  particular,  must  be  avoided  because
phthalates are commonly used  as  plasticizers  and  are easily extracted from
plastic materials.  Serious phthalate contamination  may result at any time if
consistent quality control is not practiced.

     3.5  Glassware contamination  resulting  in  analyte  degradation;   Soap
residue on glassware may cause degradation of certain analytes.  Specifically,
aldrin, heptachlor, and most organophosphorous pesticides will degrade in this
situation.  This problem is  especially  pronounced with glassware that may be
difficult to rinse  (e.g.,  500-mL  K-D  flask).    These items should be hand-
rinsed very carefully to avoid this problem.


4.0  APPARATUS AND  MATERIALS

     4.1  Refer to  the  specific  method  of  interest  for a description of the
apparatus and materials needed.


5.0  REAGENTS

     5.1  Refer to  the  specific  method  of  interest  for a description of the
solvents  needed.

     5.2  Stock standards;   Stock  solutions may  be  prepared from  pure  standard
materials or purchased  as  certified  solutions..

          5.2.1   Purgeable  stock  standards:     Prepare  stock   standards   in
     methanol  using assayed  liquids  or  gases,  as appropriate.  Because of the
     toxicity  of  some  of   the  organohalides,   primary   dilutions  of  these
     materials should  be  prepared  in a  hood.

               5.2.1.1   Place about   9.8   mL   of  methanol  in  a  10-mL  tared
          ground-glass-stoppered volumetric flask.    Allow  the  flask to  stand,
          unstoppered,  for about 10   min   or  until  all  alcohol-wetted  surfaces
          have dried.   Weigh the flask  to  the nearest 0.1 mg.

               5.2.1.2   Using a  100-uL   syringe,   immediately  add two or more
          drops of assayed reference material  to the  flask, then  reweigh.  The
          liquid  must  fall  directly   into   the  alcohol  without contacting the
          neck of the  flask.

                5.2.1.3   Reweigh,  dilute   to   volume,  stopper,   then  mix   by
           inverting the flask several  times.   Calculate the concentration  in
          micrograms per microliter  (ug/uL)  from the  net gain in  weight.   When
           compound purity is assayed to be  96%   or greater,  the  weight may  be
           used without correction  to calculate  the concentration of the stock
           standard.  Commercially  prepared stock  standards may be used at any
           concentration if they are   certified  by  the  manufacturer or by  an
           independent source.
                                   3500 - 2
                                                          Revision
                                                          Date  September 1986

-------
               5.2.1.4  Transfer the stock  standard   solution   into  a Teflon-
          sealed screw-cap bottle.   Store,  with  minimal  headspace,  at -10*C  to
          -20*C and protect from light.

               5.2.1.5  All standards  must  be   replaced  after  1  month,  or
          sooner if comparison with check standards indicates a problem.

          5.2.2  Semi volatile stock standards:     Base/neutral   and acid  stock
     standards are prepared in  methanol.   Organochlorine pesticide  standards
     are prepared in acetone.

               5.2.2.1  Stock standard solutions  should  be stored in Teflon-
          sealed  containers  at  4*C.     The   solutions  should  be  checked
          frequently for stability.   These  solutions  must be replaced  after
          six months,  or  sooner  if  comparison  with  quality control  check
          samples indicate a problem.

     5.3  Surrogate standards:  A surrogate standard  (i.e., a chemically  inert
compound not expected to occur in  an environmental sample) should be added  to
each sample, blank,  and  matrix  spike  sample   just  prior  to extraction  or
processing.  The recovery of  the  surrogate  standard  is used to monitor for
unusual matrix  effects,  gross  sample  processing  errors,  etc.   Surrogate
recovery is  evaluated  for  acceptance  by  determining  whether the measured
concentration falls within the acceptance  limits.  Recommended surrogates for
different analyte groups  follow;  however,  these  compounds,   or others that
better correspond to the analyte group,   may  be used for other analyte groups
as well.  Normally three or more standards are  added  for each analyte group.

          5.3.1  Base/neutral  and  acid  surrogate  spiking  solutions:    The
     following are recommended surrogate standards.

               Base/neutral                       Acid

               2-Fluorobiphenyl                   2-Fluorophenol
               Nitrobenzene-ds                    2,4,6-Tribromophenol
               Terphenyl-di4                      Phenol-d5

               5.3.1.1  Prepare  a  surrogate  standard  spiking  solution  in
          methanol that contains the base/neutral compounds at a concentration
          of 100 ug/mL, and  the  acid  compounds  at  200 ug/mL for water and
          sediment/soil samples  (low- and  medium-level).   For waste samples,
          the concentration should  be  500  ug/mL  for base/neutrals and 1000
          ug/mL for acids.

          5.3.2  Organochlorine  pesticide  surrogate  spiking  solution:    The
     following  are   recommended   surrogate   standards  for  Organochlorine
     pesticides.

               Organochlorine pesticides

               Dibutylchlorendate  (DBC)
               2,4,5,6-Tetrachloro-meta-xylene  (TCMX)


                                  3500 - 3
                                                         Revision      0	
                                                         Date  September 1986

-------
          5.3.2.1  Prepare a  surrogate  standard  spiking  solution at a
     concentration of 1  ug/mL  in  acetone  for  water and sediment/soil
     samples.  For waste samples,  the concentration should be 5 ug/mL.

     5.3.3  Purgeable surrogate  spiking  solution:    The  following are
recommended surrogate standards for volatile organics.

          Purgeable organics

          p-Bromof1uorobenzene
          l,2-Dichloroethane-d4
          Toluene-ds

          5.3.3.1  Prepare a surrogate spiking  solution (as described in
     Paragraph 5.2.1 or through secondary dilution of the stock standard)
     in methanol containing the surrogate standards at a concentration of
     25 ug/mL.

5.4  Matrix spike standards;    Select  five  or  more analytes from each
     analyte group for use  in  a  spiking  solution.   The following are
     recommended matrix spike standard mixtures for a few analyte groups.
     These compounds, or  others  that  better  correspond to the analyte
     group, may be used for other analyte groups as well.

     5.4.1  Base/neutral  and acid  matrix  spiking  solution:   Prepare a
spiking   solution  in  methanol  that  contains  each  of  the  following
base/neutral compounds at 100 ug/mL  and  the acid compounds at 200  ug/mL
for  water  and   sediment/soil  samples.    The  concentration  of   these
compounds should  be  five  times  higher  for waste  samples.

          Base/neutrals                      Acids

          1,2,4-Trichlorobenzene             Pentachlorophenol
          Acenaphthene                       Phenol
          2,4-Dinitrotoluene                 2-Chlorophenol
          Pyrene                             4-Chloro-3-methylphenol
          N-Ni troso-di-n-propylami ne        4-Ni trophenol
          1,4-Di chlorobenzene

      5.4.2   Organochlorine  pesticide matrix  spiking  solution:   Prepare  a
spiking  solution   in  acetone   or  methanol  that  contains  the  following
pesticides  in  the concentrations   specified  for water and sediment/soil.
The  concentration should  be five  times higher  for waste samples.

           Pesticide                          Concentration (ug/mL)

           Lindane                                      0.2
          Heptachlor                                   0.2
          Aldrin                                        0.2
          Dieldrin                                     0.5
           Endrin                                        0.5
          4,4'-DDT                                     0.5
                              3500 - 4
                                                     Revision
                                                     Date  September 1986

-------
          5.4.3  Purgeable  matrix  spiking  solution:      Prepare  a  spiking
     solution  in  methanol  that  contains   the  following  compounds  at  a
     concentration of 25 ug/tnL.

               Purgeable organics

               1,1-Dichloroethene
               Trichloroethene
               Chlorobenzene
               Toluene
               Benzene
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See  the  introductory  material  to  the  Organic  Analyte Chapter,
Section 4.1.
7.0  PROCEDURE

     7.1  Semi volatile  organic  sample  extraction;    Water,  soil/sediment,
sludge,  and!wastesamplesrequiringanalysisfor  base/neutral  and acid
extractables and/or organochlorine pesticides  must undergo solvent extraction
prior to analysis.  This  manual  contains  four  methods that may be used for
this purpose:  Method 3510; Method  3520;  Method  3540; and Method 3550.  The
method that should be used  on  a  particular sample, is highly dependent upon
the physical characteristics of  that  sample.    Therefore, review these four
methods prior to choosing one  in particular.  Appropriate surrogate standards
and, 1f necessary, matrix spiking solutions  are  added to the sample prior to
extraction for all four methods.

          7.1.1  Method 3510:  Applicable  to the extraction and concentration
     of  water-insoluble  and  slightly  water-soluble  organics  from aqueous
     samples.  A  measured  volume  of  sample  1s  solvent  extracted using a
     separatory funnel.  The extract is dried, concentrated and, if necessary,
     exchanged into a solvent compatible  with  further analysis.  Method 3520
     should be used if  an  emulsion  forms between the solvent-sample phases,
     which can not be broken up by mechanical techniques.

          7.1.2  Method 3520:  Applicable  to the extraction and concentration
     of  water-insoluble  and  slightly  water-soluble  organics  from aqueous
     samples.  A  measured  volume  of  sample  is  extracted  with an organic
     solvent  in a continuous liquid-liquid  extractor. The solvent must  have a
     density  greater  than  that  of  the  sample.    The  extract  is  dried,
     concentrated and, if necessary, exchanged  into a solvent compatible with
     further  analysis.   The  limitations  of  Method 3510 concerning solvent-
     sample phase separation do not interfere with this procedure.
                                  3500 - 5
                                                         Revision      0
                                                         Date  September 1986

-------
          7.1.3   Method 3540:   This  is   a  procedure  for extracting nonvolatile
     and semi volatile organic  compounds  from   solids  such as soils,  sludges,
     and wastes.    A  solid sample   is  mixed  with anhydrous sodium  sulfate,
     placed into  an extraction thimble or  between  two  plugs of glass wool, and
     extracted using an  appropriate  solvent   in  a  Soxhlet  extractor.  The
     extract is   dried,  concentrated and,  if necessary,  exchanged  into a
     solvent compatible with further analysis.

          7.1.4   Method 3550:   This  method  is  applicable  to the  extraction of
     nonvolatile  and semi volatile organic  compounds  from solids such as  soils,
     sludges,  and wastes using  the   technique  of sonication.  Two procedures
     are detailed depending upon the expected concentration of organics  in the
     sample; a low concentration and a  high concentration method.   In both, a
     known weight of sample is mixed with  anhydrous  sodium sulfate and solvent
     extracted using sonication.  The extract   is dried,  concentrated and, if
     necessary,  exchanged into a solvent compatible  with further  analysis.

          7.1.5   Method 3580:   This  method  describes  the  technique of solvent
     dilution of non-aqueous waste samples.     It  is  designed for wastes that
     may contain  organic chemicals at a  level greater than 20,000 mg/kg and
     that are soluble in the dilution solvent.

     7.2  Volatile organic sample preparation;     There  are  three methods for
volatile sample preparation:  Method  5030;  Method 5040; and  direct  injection.
Method 5030 is the most  widely  applicable  procedure  for  analysis of  volatile
organics, while the direct injection  technique may  have  limited  applicability
to aqueous matrices.

          7.2.1  Method 5030:   This  method   describes the technique  of  purge-
     and-trap  for  the  introduction   of   purgeable  organics   into  a  gas
     chromatograph.  This procedure is applicable  for use  with  aqueous samples
     directly  and  to  solids,  wastes,  soils/sediments,  and  water-miscible
     liquids following  appropriate  preparation.     An  inert   gas  is bubbled
     through  the   sample,  which  will   efficiently  transfer  the   purgeable
     organics from  the aqueous phase to  the  vapor phase.  The  vapor phase  is
     swept  through  a sorbent  trap  where  the  purgeables are trapped.   After
     purging is completed, the trap  is  heated and backflushed  with the inert
     gas to desorb  the purgeables onto a gas chromatographic column.   Prior to
     application  of   the  purge-and-trap  procedure,   all  samples (including
     blanks, spikes, and duplicates) should be spiked with surrogate standards
     and,  if required, with matrix  spiking compounds.

           7.2.2  Method  5040:   This method  is applicable to the investigation
     of sorbent cartridges  from  volatile organic  sampling train  (VOST).

     7.3   Sample analysis;  Following preparation   of  a  sample by one of the
methods described  above,   the  sample   is   ready  for  further   analysis.  For
samples requiring  volatile  organic  analysis,   application  of one of  the three
methods described  above  is  followed  directly by gas chromatographic analysis
(Methods 8010,  8015,   8020,   or 8030).    Samples prepared for semivolatile
analysis may,  if   necessary,   undergo  cleanup   (See  Method  3600)  prior to
application of a specific  determinative method.


                                  3500 -  6
                                                          Revision       0
                                                         Date  September 1986

-------
8.0  QUALITY CONTROL

     8.1  Refer to  Chapter  One  for  specific  guidance  on  quality control
procedures.

     8.2  Before  processing  any  samples,  the  analyst  should  demonstrate
through the analysis of a reagent  water blank that all glassware and reagents
are interference free.  Each  time  a  set  of samples are processed, a method
blank(s)  should  be  processed  as  a  safeguard  against  chronic laboratory
contamination.  The blank samples should  be carried through all stages of the
sample preparation and measurement.

     8.3  Surrogate standards should be added to all samples when specified in
the appropriate determinative method in Chapter Four, Section 4.3.

     8.4  A reagent blank, a  matrix  spike,  and  a duplicate or matrix spike
duplicate must be performed for each  analytical  batch  (up to a maximum of 20
samples) analyzed.

     8.5  For GC or GC/MS analysis,  the analytical system performance must be
verified by  analyzing  quality  control   (QC)  check  samples.   Method 8000,
Section  8.0  discusses  in  detail  the  process  of  verification;  however,
preparation of the QC check  sample  concentrate  is dependent upon the method
being evaluated.

          8.5.1  Volatile  organic  QC  check   samples:     QC  check  sample
     concentrates containing each analyte of  interest are spiked into reagent
     water  (defined as the  QC  check  sample)  and analyzed by purge-and-trap
      (Method 5030).  The concentration of  each analyte  in the QC check sample
      is 20  ug/L.  The evaluation  of system performance  is discussed  in detail
      in Method 8000, beginning with Paragraph 8.6.

          8.5.2  Semivolatile organic  QC  check  samples:    To  evaluate the
     performance of  the  analytical  method,  the  QC   check  samples must be
      handled in exactly the same manner  as actual  samples.  Therefore, 1.0 mL
     of the QC  check  sample  concentrate  is  spiked   into  each of four 1-L
      aliquots of reagent water   (now  called  the QC check sample),  extracted,
      and then analyzed by GC.   The variety of semi volatile  analytes  which may
      be analyzed by GC is such  that  the  concentration  of the QC check sample
      concentrate   is  different    for   the  different   analytical   techniques
      presented  in  the manual.   Method  8000 discusses  in detail the  procedure
      of verifying  the  detection   system   once  the QC  check  sample has been
      prepared.  The concentrations of the  QC check sample concentrate  for the
      various methods  are  as  follows:

                8.5.2.1  Method  8040  -    Phenols;       The  QC  check  sample
           concentrate  should  contain each  analyte  at a concentration of
           100  ug/mL in 2-propanol.

                8.5.2.2  Method  8060  - Phthalate   esters;  The  QC  check sample
           concentrate  should  contain thefollowinganalytes  at the following
           concentrations  in  acetone:  butyl benzyl  phthalate,  10 ug/mL; bis(2-
           ethylhexyl)phthalate,  50 ug/mL;   di-n-octylphthalate,  50  ug/mL; and
           any other phthalate at  25  ug/mL.

                                   3500 -  7
                                                          Revision       0
                                                          Date  September 1986

-------
              8.5.2.3  Method 8080  - Organochlorine pesticides and PCBs:  The
          QC  check   sampTeconcentrate   shouldcontain each single-component
          analyte at the  following   concentrations  in  acetone:  4,4'-DDD,  10
          ug/mL; 4,4'-DDT,   10   ug/mL;   endosulfan  II,  10  ug/mL; endosulfan
          sulfate,  10 ug/mL;  and  any   other  single-component pesticide at 2
          ug/mL.  If the  method  is only  to  be  used  to analyze PCBs, chlordane,
          or  toxaphene, the  QC  check   sample  concentrate  should  contain the
          most representative multicomponent   parameter  at  a concentration  of
          50  ug/mL  in acetone.

               8.5.2.4 Method 8090  -  Nitroaromatics  and  Cyclic  Ketones:  The
          QC  check   sampleconcentrateshouldcontaineachanalyte at the
          following concentrations  in  acetone:  each  dinitrotoluene at
          20  ug/mL; and isophorone  and nitrobenzene at  100 ug/mL.

               8.5.2.5  Method 8100  -  Polynuclear   aromatic hydrocarbons:  The
          QC  check   sampleconcentrateshouldcontaineachanalyte at the
          following concentrations  in   acetonitrile:    naphthalene, 100 ug/mL;
          acenaphthylene, 100 ug/mL; acenaphthene,  100  ug/mL;  fluorene,
          100  ug/mL;  phenanthrene,   100   ug/mL;   anthracene,   100  ug/mL;
          benzo(k)fluoranthene  5 ug/mL;  and any  other PAH  at 10 ug/mL.

               8.5.2.6  Method  8120 -  Chlorinated   hydrocarbons;   The QC check
          sample concentrate  should  contain  each  analyte  at the following
          concentrations in acetone:  hexachloro-substituted hydrocarbons,
          10 ug/mL; and any other chlorinated hydrocarbon, 100 ug/mL.


9.0  METHOD PERFORMANCE

     9.1  The recovery  of  surrogate  standards  is   used  to monitor unusual
          matrix effects, sample processing  problems,   etc.   The recovery  of
          matrix  spiking  compounds  indicates  the  presence  or  absence  of
          unusual  matrix effects.

     9.2  The performance of  this  method  will  be  dictated  by the overall
          performance  of  the  sample  preparation  in  combination  with  the
          analytical determinative method.


10.0  REFERENCES

     10.1  None required.
                                   3500 - 8
                                                          Revision
                                                          Date   September  1986

-------
                                                 L iJ

                                                 5F
F Z
111 O
X 
-------
3.3  METHODS FOR DETERMINATION OF METALS

          This  section of the manual  contains  seven  analytical  techniques for
trace metal  determinations:  inductively coupled argon  plasma atomic emission
spectrometry (ICP-AES), inductively coupled plasma mass spectrometry (ICP-MS),
direct-aspiration  or  flame  atomic  absorption  spectrometry  (FLAA),  graphite-
furnace  atomic  absorption  spectrometry  (GFAA),   hydride-generation  atomic
absorption spectrometry (HGAA), cold-vapor atomic absorption spectrometry (CVAA),
and  several  procedures for  hexavalent  chromium analysis.   Each of  these is
briefly discussed below in terms of  advantages, disadvantages, and cautions for
analysis of wastes.

          ICP's primary  advantage  is that  it allows  simultaneous or  rapid
sequential  determination  of many  elements  in  a  short  time.   The  primary
disadvantage of ICP is background radiation from other elements and the plasma
gases.  Although all  ICP  instruments  utilize  high-resolution optics and back-
ground correction to minimize these interferences,  analysis for traces of metals
in the presence of  a large excess of a single metal  is difficult. Examples would
be traces of metals in  an alloy or  traces  of metals  in a limed (high calcium)
waste.  ICP and Flame  AA have comparable detection limits (within  a factor of 4)
except that  ICP exhibits  greater sensitivity for refractories (Al,  Ba,  etc.).
Furnace AA, in general, will exhibit lower detection  limits than either ICP or
FLAA. Detection limits are drastically improved when ICP-MS is used.  In general
ICP-MS exhibits greater sensitivity  than either GFAA of  FLAA for most elements.
The greatest disadvantage  of ICP-MS  is isobaric elemental  interferences.  These
are caused by different elements  forming atomic ions with the same nominal mass-
to-charge ratio.  Mathematical correction for interfering ions can minimize these
interferences.

          Flame AAS (FLAA) direct aspiration determinations, as opposed to ICP,
are normally completed  as  single element analyses  and  are  relatively  free of
interelement  spectral  interferences.    Either a  nitrous-oxide/acetylene  or
air/acetylene flame is used as an energy source for dissociating the aspirated
sample into the free atomic  state making  analyte atoms available for absorption
of light.   In the analysis  of some elements the temperature or type of flame used
is critical.   If  the proper flame  and analytical  conditions  are not  used,
chemical  and ionization interferences can occur.

          Graphite  Furnace AAS (GFAA) replaces the flame  with an electrically
heated graphite furnace.  The furnace allows  for gradual heating of the sample
aliquot  in  several   stages.    Thus,  the  processes  of  desolvation,  drying,
decomposition of organic  and inorganic  molecules  and salts,  and  formation of
atoms which must  occur in a flame or ICP in  a few milliseconds may be allowed to
occur over  a much  longer time  period and at  controlled temperatures  in  the
furnace.  This allows an experienced analyst to remove unwanted matrix components
by using  temperature programming and/or matrix  modifiers.   The major advantage
of this  technique is that  it  affords extremely low detection limits.  It is the
easiest to perform on relatively  clean  samples.  Because  this technique  is so
sensitive, interferences can be  a real problem; finding the optimum combination
of digestion,  heating times and  temperatures, and matrix modifiers can  be  a

                                  THREE  - 7                       Revision 2
                                                                  September  1994

-------
challenge for complex matrices.

          Hydride AA utilizes a chemical reduction to reduce and separate arsenic
or selenium selectively from a sample digestate.  The technique therefore has the
advantage of being  able to isolate these two elements from complex samples which
may  cause   interferences  for  other  analytical   procedures.     Significant
interferences have  been reported when any of the following is present:   1) easily
reduced metals (Cu, Ag, Hg);  2) high concentrations of transition metals (>200
mg/L); 3)  oxidizing  agents  (oxides  of  nitrogen)  remaining following  sample
digestion.

          Cold-Vapor AA uses a chemical reduction  to  reduce mercury selectively.
The procedure is extremely sensitive but is subject to interferences from some
volatile organics,  chlorine,  and sulfur compounds.
                                   THREE -  8                      Revision 2
                                                                  September 1994

-------
                                 METHOD 6010A

            INDUCTIVELY  COUPLED PLASMA-ATOM1C  EMISSION  SPECTROSCOPY
1.0   SCOPE AND APPLICATION

      1.1   Inductively  coupled  plasma-atomic  emission  spectroscopy  (ICP)
determines  trace  elements,   including  metals,  in  solution.  The  method  is
applicable to  all  of the elements listed in Table  1.  All  matrices, including
ground water,  aqueous samples, TCLP  and  EP extracts,  industrial  and organic
wastes, soils,  sludges,  sediments,  and other  solid  wastes,  require digestion
prior to analysis.

      1.2   Elements for which Method 6010  is applicable are listed  in Table 1.
Detection limits, sensitivity, and optimum ranges of the metals will vary with
the matrices  and model  of spectrometer.  The  data  shown  in  Table  1  provide
estimated  detection   limits   for  clean  aqueous   samples   using  pneumatic
nebulization.  Use of  this method  is restricted  to spectroscopists who  are
knowledgeable   in  the  correction   of   spectral,  chemical,   and  physical
interferences.
2.0   SUMMARY OF METHOD

      2.1   Prior to  analysis,  samples must be solubilized  or digested using
appropriate Sample Preparation Methods  (e.g. Methods 3005-3050). When analyzing
for dissolved constituents, acid digestion is not necessary if the samples are
filtered and acid preserved prior to analysis.

      2.2   Method 6010 describes the simultaneous, or sequential, multielemental
determination of elements by ICP. The method measures element-emitted light by
optical  spectrometry.    Samples are  nebulized  and  the  resulting  aerosol  is
transported to the plasma torch. Element-specific atomic-line emission spectra
are produced by a radio-frequency inductively coupled plasma.  The spectra are
dispersed  by  a grating  spectrometer,  and  the  intensities  of the  lines  are
monitored by photomultiplier tubes.  Background correction  is required for trace
element determination. Background must be measured  adjacent to  analyte lines on
samples during  analysis.   The  position  selected  for the background-intensity
measurement, on either or both  sides of the analytical line, will be determined
by the complexity of  the  spectrum adjacent  to the  analyte line.   The position
used must  be free  of spectral  interference  and  reflect  the same  change  in
background intensity as occurs  at the analyte wavelength measured.  Background
correction  is  not  required  in  cases  of line  broadening where  a background
correction  measurement would  actually  degrade  the  analytical  result.   The
possibility of  additional  interferences  named  in  Section 3.0 should  also  be
recognized  and  appropriate  corrections  made;  tests for  their  presence  are
described in Step 8.5.
                                   6010A  -  1                       Revision 1
                                                                  July 1992

-------
                                   TABLE 1.
     RECOMMENDED WAVELENGTHS AND ESTIMATED INSTRUMENTAL DETECTION LIMITS
Detection
Element
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Lithium
Magnesium
Manganese
Molybdenum
Nickel
Phosphorus
Potassium
Selenium
Silver
Sodium
Strontium
Thallium
Vanadium
Zinc
Wavelength8 (nm)
308.215
206.833
193.696
455.403
313.042
226.502
317.933
267.716
228.616
324.754
259.940
220.353
670.784
279.079
257.610
202.030
231.604
213.618
766.491
196.026
328.068
588.995
407.771
190.864
292.402
213.856
Estimated
Limit (ug/L)
45
32
53
2
0.3
4
10
7
7
6
7
42
5
30
2
8
15
51
See note c
75
7
29
0.3
40
8
2
                                                                                   i
           wavelengths listed are recommended because of their sensitivity and
overall acceptance. Other wavelengths may be substituted if they can provide
the needed sensitivity and are treated with the same corrective techniques for
spectral interference (see Step 3.1). In time, other elements may be added as
more information becomes available and as required.
           estimated instrumental detection limits shown are taken from
Reference 1 in Section 10.0 below. They are given as a guide for an
instrumental limit. The actual method detection limits are sample dependent
and may vary as the sample matrix varies.
      Highly dependent on operating conditions and plasma position.


                                                                  Re
                                                                  July 1992
6010A - 2                       Revision  1       JB

-------
3.0   INTERFERENCES

      3.1  Spectral  interferences are caused by:  (1)  overlap of a spectral line
from another element at the analytical or background measurement wavelengths; (2)
unresolved overlap of molecular band spectra; (3) background contribution from
continuum or recombination phenomena; and (4) stray light  from the line emission
of high-concentration  elements.   Spectral  overlap  can be compensated  for by
computer-correcting the raw data  after monitoring  and measuring the interfering
element.  Unresolved  overlap  requires  selection  of  an  alternate  wavelength.
Background  contribution  and stray light can  usually be  compensated  for by  a
background correction adjacent to the analyte line.

      Users  of  all   ICP   instruments  must  verify   the  absence  of  spectral
interference  from  an element  in  a  sample  for which  there  is no  instrument
detection channel. Recommended wavelengths are listed in  Table 1  and potential
spectral interferences for the recommended wavelengths are given in Table  2. The
data in  Table 2 are  intended  as  rudimentary guides for  indicating potential
interferences;  for  this purpose, linear  relations  between concentration and
intensity for the analytes and the interferents can  be assumed.

            3.1.1  Element-specific  interference  is  expressed   as   analyte
      concentration equivalents (i.e. false analyte concentrations) arising from
      100 mg/L of the  interference element. For example, assume that As is to be
      determined (at 193.696 nm) in a sample containing  approximately 10 mg/L of
      Al. According to Table 2, 100 mg/L of Al would yield a false signal  for As
      equivalent to approximately 1.3 mg/L. Therefore,  the presence of 10 mg/L
      of Al would result  in  a  false  signal  for As equivalent  to  approximately
      0.13  mg/L.  The  user is cautioned  that  other instruments  may exhibit
      somewhat different levels of interference than  those shown in Table  2. The
      interference effects must  be  evaluated  for each  individual  instrument
      since the intensities will  vary with operating conditions, power, viewing
      height,  argon flow rate, etc.  The user should be  aware of the possibility
      of interferences other than those specified in Table 2 and  that analysts
      should be aware of these interferences when conducting analyses.

            3.1.2  The  dashes   in   Table   2   indicate  that  no   measurable
      interferences were  observed  even at  higher interferent  concentrations.
      Generally,  interferences  were discernible  if they  produced peaks,  or
      background shifts, corresponding  to  2  to  5%  of the peaks generated  by the
      analyte concentrations.

            3.1.3  At present, information  on  the listed silver  and potassium
      wavelengths is not available,  but it has been  reported that  second-order
      energy from the magnesium 383.231-nm wavelength  interferes with the listed
      potassium line at 766.491 nm.
                                   6010A  - 3                       Revision 1
                                                                  July 1992

-------
                                     TABLE  2.

                  ANALYTE  CONCENTRATION  EQUIVALENTS ARISING  FROM
                        INTERFERENCE  AT  THE 100-mg/L  LEVEL
                                                Interferent
                                                           a,b
 Analyte
Wavelength  --•

   (run)     Al
Ca    Cr    Cu    Fe
         Mg    Mn
N1
Tl
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Selenium
Sodium
Thallium
Vanadium
Zinc
308.215
206.833
193.696
455.403
313.042
226.502
317.933
267.716
228.616
324.754
259.940
220.353
279.079
257.610
202.030
231.604
196.026
588.995
190.864
292.402
213.856

0.47 --
1.3 --
	 	
__
	 	
__

__
--
	 —
0.17 --
0.02
0.005 --
0.05 --
__
0.23 --
__
0.30 --
__
- -

2.9
0.44
_ _
--
_ _
0.08
--
0.03
--
—
--
0.11
0.01
—
--
--

--
0.05
--
0.21
0.08 --
--
	 _ — __ __
--
0.03 --
0.01 0.01 0.04
0.003 -- 0.04
0.005 --
0.003 --
	 0.12
__
0.13 -- 0.25
0.002 0.002 --
0.03 --
_.
0.09 --
__
__
0.005 --
0.14 - 	

0.25
--
_. _ _ _
0.04
0.02 --
0.03
__
0.03 0.15
0.05
— —
_.
0.07
— —
_ _
__
__
0.08
_.
0.02
0.29 --
1.4
0.45
1.1
	
0.05
	
0.03
0.04
--
0.02
—
--
0.12
—

--
--
--
--
--

Cashes indicate that no interference was observed even when interferents were
introduced at the following levels:
      Al  -
      Ca -
      Cr -
      Cu -
      Fe -
1000 mg/L
1000 mg/L
 200 mg/L
 200 mg/L
1000 mg/L
         Mg
         Mn
         Tl
          V
1000 mg/L
 200 mg/L
 200 mg/L
 200 mg/L
     figures recorded as analyte concentrations are not the actual observed
concentrations; to obtain those figures, add the listed concentration to the
interferent figure.
                                   6010A  -  4
                                                      Revision  1
                                                      July  1992

-------
      3.2   Physical  interferences  are  effects  associated with  the  sample
nebulization and transport processes. Changes in viscosity and surface tension
can  cause significant  inaccuracies,  especially  in  samples containing  high
dissolved solids or high acid concentrations.  Differences  in solution volatility
can  also  cause inaccuracies when  organic  solvents  are  involved.  If physical
interferences  are  present,  they  must be reduced by diluting the  sample  or by
using a peristaltic pump.  Another problem that can occur with  high  dissolved
solids is salt buildup at the tip of the nebulizer,  which affects aerosol flow
rate and causes instrumental drift. The problem can be  controlled by wetting the
argon  prior  to  nebulization,  using  a tip  washer,   or  diluting the  sample.
Changing the nebulizer and removing salt buildup at the tip of the torch sample
injector can be used as an additional measure to control salt  buildup.  Also, it
has been reported that better control of the argon  flow rate improves instrument
performance; this  is accomplished with the use of mass flow controllers.

      3.3   Chemical  interferences   include   molecular   compound  formation,
ionization effects, and solute vaporization effects.  Normally, these effects are
not significant with the  ICP technique. If observed,  they  can  be minimized by
careful selection of operating conditions (incident power, observation position,
and so forth), by  buffering of the sample, by matrix matching,  and by standard
addition procedures. Chemical  interferences are highly dependent on matrix type
and the specific analyte element.


4.0  APPARATUS AND MATERIALS

      4.1  Inductively coupled argon plasma emission spectrometer:

            4.1.1   Computer-controlled emission  spectrometer with  background
      correction.

            4.1.2   Radio frequency generator compliant with FCC regulations.

            4.1.3  Argon gas supply  - Welding grade or better.

      4.2  Operating conditions  - The analyst should follow the instructions
provided by the instrument manufacturer. For operation with organic solvents, use
of the auxiliary  argon  inlet  is  recommended,  as  are solvent-resistant tubing,
increased plasma (coolant) argon flow, decreased nebulizer flow, and increased
RF  power  to  obtain stable operation  and precise measurements.  Sensitivity,
instrumental detection limit,  precision, linear dynamic range, and interference
effects must be established for each  individual analyte line on that particular
instrument. All measurements must  be within  the  instrument  linear range where
spectral interference correction  factors are  valid. The analyst must (1) verify
that the instrument configuration and operating conditions satisfy the analytical
requirements  and  (2)  maintain  quality  control  data   confirming  instrument
performance and analytical results.

      4.3  Class A  volumetric flasks

      4.4  Class A  volumetric pipets
                                   6010A  -  5                       Revision 1
                                                                  July 1992

-------
      4.5   Analytical balance - capable of accurate measurement to 4 significant
figures.


5.0  REAGENTS

      5.1  Reagent grade chemicals shall  be  used  in all tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform to the specifications
of the Committee on Analytical Reagents  of the American Chemical Society, where
such specifications are available. Other grades may be used,  provided it is first
ascertained that the  reagent  is  of sufficiently  high  purity to permit its use
without lessening the accuracy of the determination. If the  purity of a reagent
is in question analyze for contamination.  If the  concentration is less than the
MDL then the reagent  is acceptable.

            5.1.1  Hydrochloric acid  (cone), HC1.

            5.1.2  Hydrochloric acid (1:1),  HC1.  Add 500 ml  concentrated HC1 to
      400 ml water and dilute to 1 liter  in an appropriate beaker.

            5.1.3  Nitric acid (cone), HN03.

            5.1.4  Nitric acid (1:1), HN03. Add 500  ml concentrated  HN03 to
      400 ml water and dilute to 1 liter  in an appropriate beaker.

      5.2   Reagent Water. All references to water in the method refer to reagent
water unless otherwise specified.  Reagent water will  be interference free.
Refer to Chapter One  for  a definition of  reagent water.

      5.3   Standard  stock solutions may be purchased or prepared from ultra-
high purity grade chemicals or metals (99.99 to 99.999% pure).  All  salts must be
dried for 1 hour at 105°C, unless otherwise specified.

      CAUTION:  Many  metal salts are extremely  toxic  if inhaled or swallowed.
                Wash  hands thoroughly after handling.

      Typical stock solution  preparation  procedures follow. Concentrations are
calculated based upon the weight of pure metal  added, or with the use of the mole
fraction and the weight of the metal  salt added.

Metal

          Concentration  (ppm)  -

Metal salts

          Concentration  (pp.)  -


            5.3.1  Aluminum solution, stock, 1 ml = 1000 ug Al:  Dissolve  1.0 g
      of  aluminum  metal, weighed   accurately to  at  least  four  significant
      figures,  in  an  acid mixture of 4 ml of (1:1)  HC1  and  1 ml  of concentrated
      HN03  in  a  beaker. Warm gently  to effect  solution.   When  solution is
      complete,  transfer quantitatively  to  a liter  flask,  add an additional

                                   6010A - 6                      Revision 1
                                                                  July 1992

-------
10 ml of  (1:1)  HC1  and dilute to volume in  a  1,000  ml volumetric flask
with water.

      5.3.2  Antimony  solution,  stock,  1  ml  =  1000  ug  Sb:   Dissolve
2.70 g K(SbO)C4H,06  (mole fraction Sb = 0.3749), weighed accurately to at
least four significant figures,  in water, add 10 mL (1:1) HC1, and dilute
to volume in a 1,000 ml volumetric flask with water.

      5.3.3  Arsenic solution, stock, 1 mL = 1000 ug As: Dissolve 1.30 g
of ASp03  (mole fraction As = 0.7574), weighed accurately to at least four
significant figures, in 100 mL of water containing 0.4 g NaOH. Acidify the
solution with 2 mL  concentrated HN03 and dilute to volume in  a 1,000 mL
volumetric flask with water

      5.3.4  Barium solution, stock, 1 mL = 1000 ug Ba:  Dissolve 1.50 g
BaCl2 (mole fraction  Ba  =  0.6595), dried at 250°C for 2  hours, weighed
accurately to at least four significant figures, in 10 mL water with  1 mL
(1:1) HC1.  Add  10.0  mL  (1:1) HC1  and dilute to  volume  in a  1,000 mL
volumetric flask with water.

      5.3.5  Beryllium solution, stock,  1 mL = 1000  ug Be:   Do not dry.
Dissolve 19.7 g  BeS04'4H20 (mole  fraction  Be = 0.0509),  weighed accurately
to at least four significant figures, in water, add 10.0 mL concentrated
HN03, and dilute to volume  in a  1,000 mL volumetric flask with   water.

      5.3.6  Cadmium solution, stock, 1 mL  =  1000  ug  Cd:  Dissolve 1.10 g
CdO  (mole  fraction Cd  =  0.8754), weighed  accurately to at  least  four
significant figures, in a minimum amount of  (1:1) HN03. Heat to increase
rate of dissolution. Add 10.0 mL concentrated HN03 and dilute  to  volume in
a 1,000 mL volumetric flask with water.

      5.3.7  Calcium solution, stock, 1 mL = 1000 ug Ca:  Suspend 2.50 g
CaCO, (mole  Ca  fraction =  0.4005),  dried  at  180°C  for  1  hour  before
weighing,  weighed  accurately to  at  least  four  significant  figures, in
water and dissolve  cautiously with a minimum amount of  (1:1) HN03. Add 10.0
mL concentrated HN03 and dilute  to volume in a 1,000 mL volumetric flask
with water.

      5.3.8  Chromium  solution,  stock,  1  mL  =  1000  ug  Cr:   Dissolve
1.90 g CrO, (mole fraction Cr =  0.5200), weighed accurately to at least
four significant figures,  in water. When  solution  is complete, acidify
with 10 mL concentrated HN03 and dilute to volume in a 1,000 mL volumetric
flask with water.

      5.3.9  Cobalt solution, stock, 1 mL = 1000  ug Co: Dissolve  1.00 g
of cobalt metal, weighed accurately to at least four significant figures,
in a minimum  amount of (1:1) HNO,. Add 10.0 mL  (1:1)  HC1  and   dilute to
volume in a 1,000 mL volumetric flask with water.

      5.3.10  Copper solution, stock, 1 mL = 1000 ug Cu: Dissolve 1.30 g
CuO  (mole  fraction Cu  =  0.7989), weighed  accurately to at  least  four
significant figures),  in a  minimum amount  of (1:1)  HN03.   Add 10.0 mL


                             6010A -  7                      Revision 1
                                                            July 1992

-------
concentrated HN03 and dilute to volume in a 1,000 mL volumetric flask with
water.

      5.3.11   Iron solution,  stock,  1 ml  = 1000 ug  Fe:   Dissolve 1.40 g
Fe203  (mole fraction Fe  =  0.6994),  weighed accurately to  at least four
significant figures,  in  a warm mixture of 20 mL (1:1) HC1  and  2 ml of
concentrated HN03. Cool, add an additional  5.0 ml of concentrated HN03, and
dilute to volume in a 1,000 ml volumetric  flask with water.

      5.3.12   Lead solution,  stock,  1 mL  = 1000 ug  Pb:   Dissolve 1.60 g
Pb(NO,)2  (mole  fraction Pb  = 0.6256), weighed  accurately to at least four
significant figures, in  a  minimum amount  of (1:1) HN03.  Add 10 mL (1:1)
HN03 and dilute to volume in a 1,000 mL  volumetric flask with water.

      5.3.13  Lithium solution,  stock, 1 mL = 1000 ug  Li: Dissolve  5.324 g
lithium carbonate (mole  fraction  Li  = 0.1878),  weighed accurately to at
least four significant figures,  in  a minimum  amount of  (1:1)  HC1  and
dilute to volume in a 1,000 mL volumetric  flask with water.

      5.3.14   Magnesium solution,  stock,  1  mL  = 1000 ug  Mg:  Dissolve
1.70 g MgO  (mole fraction Mg = 0.6030), weighed  accurately to at least
four significant figures, in a minimum amount of (1:1) HN03.  Add   10.0 mL
(1:1) concentrated HN03 and dilute to volume in a 1,000 mL volumetric flask
with water.

      5.3.15   Manganese  solution, stock,  1  mL  = 1000 ug  Mn:   Dissolve
1.00 g of manganese metal,  weighed accurately to at least four significant
figures,  in  acid mixture (10 mL  concentrated HC1  and 1  mL  concentrated
HN03)  and dilute to volume in a 1,000 mL volumetric flask with water.

      5.3.16   Molybdenum solution,  stock, 1 mL  =  1000  ug Mo:   Dissolve
2.00 g (NH,)6Mo7024.4HpO (mole fraction Mo = 0.5772), weighed  accurately to
at  least  four  significant figures,  in  water and dilute  to  volume in a
1,000 mL  volumetric flask with water.

      5.3.17  Nickel  solution, stock, 1 mL  = 1000 ug Mi:  Dissolve   1.00 g
of nickel metal, weighed accurately to at  least  four  significant figures,
in 10.0 mL hot  concentrated HN03,  cool, and dilute to volume in  a 1,000 mL
volumetric flask with water.

      5.3.18   Phosphate  solution,  stock,  1  mL  =  1000  ug  P:  Dissolve
4.393 g anhydrous KH?P04  (mole fraction P = 0.2276), weighed  accurately to
at least  four significant figures, in water.  Dilute  to volume  in  a 1,000
mL volumetric  flask with water.

      5.3.19  Potassium solution,  stock, 1  mL = 1000 ug K:  Dissolve 1.90 g
KC1  (mole fraction K = 0.5244) dried at 110°C,  weighed accurately to at
least four significant figures, in water,  and dilute  to volume  in  a 1,000
mL volumetric  flask with water.

      5.3.20   Selenium solution,  stock,  1  mL = 1000 ug Se:  Do not  dry.
Dissolve  1.70  g  H2Se03 (mole fraction Se =  0.6123), weighed  accurately to


                             6010A - 8                       Revision 1
                                                             July  1992

-------
      at least four  significant figures, in water  and  dilute to volume  in  a
      1,000 mL volumetric flask with water.

            5.3.21  Silver solution, stock,  1 ml = 1000  ug Ag:  Dissolve
      1.60 g AgNO, (mole fraction Ag = 0.6350),  weighed  accurately to at least
      four significant figures,  in water and  10  ml concentrated HN03. Dilute to
      volume in a 1,000 mL volumetric flask with water.

            5.3.22  Sodium solution, stock,  1 ml = 1000 ug Na:  Dissolve 2.50 g
      NaCl  (mole  fraction  Na =  0.3934), weighed  accurately to at  least  four
      significant figures,  in water. Add  10.0 ml concentrated HN03 and dilute to
      volume in a 1,000 ml volumetric flask with water.

            5.3.23   Strontium  solution, stock,  1  ml =  1000 ug  Sr:  Dissolve
      2.415 g of  strontium  nitrate (Sr(NO,)2) (mole  fraction  0.4140),  weighed
      accurately  to  at  least  four  significant figures,  in  a 1-liter  flask
      containing 10 ml of concentrated HC1 and 700 mL of water. Dilute to volume
      in a 1,000 mL volumetric  flask with water.

            5.3.24   Thallium solution,  stock,  1 mL =  1000  ug  Tl:  Dissolve
      1.30 g T1NO, (mole fraction Tl = 0.7672),  weighed  accurately to at least
      four  significant  figures,  in water. Add  10.0 mL  concentrated  HN03 and
      dilute to volume in a 1,000 mL volumetric flask with water.

            5.3.25  Vanadium solution, stock, 1 mL =  1000 ug V: Dissolve 2.30 g
      NH403  (mole fraction  V =  0.4356), weighed  accurately to at  least  four
      significant figures,  in  a minimum amount of  concentrated  HN03.  Heat to
      increase rate of dissolution. Add  10.0  mL  concentrated HN03 ana dilute to
      volume in a 1,000 mL volumetric flask with water.

            5.3.26  Zinc solution,  stock, 1  mL  =  1000 ug Zn:   Dissolve 1.20 g
      ZnO  (mole  fraction Zn =  0.8034), weighed  accurately to at  least  four
      significant figures,  in  a  minimum amount of  dilute HNO,.  Add  10.0 mL
      concentrated HN03 and dilute  to volume in a 1,000 mL volumetric flask with
      water.

      5.4    Mixed calibration  standard  solutions -  Prepare mixed calibration
standard solutions by  combining  appropriate  volumes  of  the stock solutions in
volumetric flasks (see Table 3).   Matrix match  with the appropriate acids and
dilute to 100 mL with water.  Prior to  preparing  the mixed standards, each stock
solution  should  be   analyzed   separately   to   determine  possible  spectral
interference or the presence of  impurities. Care should be taken when preparing
the mixed  standards  to  ensure   that  the elements  are  compatible  and  stable
together. Transfer the mixed standard solutions to FEP fluorocarbon or previously
unused polyethylene or polypropylene bottles  for storage. Fresh mixed standards
should be prepared, as needed, with the realization that concentration can change
on aging.  Calibration standards  must be initially  verified using  a  quality
control sample (see Step 5.8) and monitored weekly for stability.  Some typical
calibration standard combinations are listed in Table 3. All  mixtures should then
be scanned using  a sequential spectrometer to verify  the absence of interelement
spectral interference in the recommended mixed standard  solutions.
                                   6010A  -  9                       Revision 1
                                                                  July 1992

-------
      NOTE:  If the  addition  of  silver  to  the  recommended  acid  combination
            results in an  initial precipitation, add 15 ml of water and warm the
            flask until the solution clears.  Cool  and dilute  to 100 mL
            with  water.  For this acid  combination,  the silver  concentration
            should be  limited  to 2 mg/L.  Silver under  these conditions is stable
            in a  tap-water matrix for 30 days. Higher concentrations of silver
            require additional HC1.
                                   TABLE 3.
                           MIXED STANDARD SOLUTIONS
      Solution                      Elements
         I                    Be, Cd, Mn, Pb,  Se and Zn
         II                   Ba, Co, Cu, Fe,  and V
         III                  As, Mo
         IV                   Al, Ca, Cr, K, Na, Ni,Li,& Sr
         V                    Ag (see Note to Step 5.4), Mg, Sb, and Tl
         VI                   P
      5.5  Two types of  blanks  are  required  for the analysis.  The calibration
blank is used  in establishing the analytical curve, and the reagent blank is used
to correct for possible contamination resulting from varying amounts of the acids
used in the sample processing.

            5.5.1  The calibration blank is prepared by acidifying reagent water
      to  the  same  concentrations of  the acids  found  in  the  standards  and
      samples.   Prepare  a  sufficient quantity to  flush  the  system  between
      standards and samples.

            5.5.2  The method blank  must  contain all the  reagents and  in the
      same volumes as used  in the processing of the samples.  The method blank
      must be carried through the complete procedure and  contain  the  same acid
      concentration  in  the  final solution  as  the  sample solution used  for
      analysis.

      5.6  The instrument check  standard is prepared by the  analyst by combining
compatible  elements  at  concentrations equivalent  to the  midpoint of  their
respective calibration curves (see Step 8.6.1.1  for  use).  The instrument check
standard  should  be prepared from a  source  independent from that used  in the
calibration standards.

      5.7   The  interference  check  solution   is  prepared  to   contain  known
concentrations of interfering elements  that will provide an adequate test of the
correction factors. Spike the sample with the elements of interest at approximate

                                  6010A - 10                       Revision 1
                                                                   July 1992

-------
      5.7   The  interference  check  solution  is  prepared  to  contain  known
concentrations of interfering elements that will provide an  adequate test of the
correction factors.  Spike the sample with the elements of interest at approximate
concentrations of 10 times  the instrumental detection limits. In the absence of
measurable analyte,  overcorrection could go undetected because a negative value
could  be  reported  as  zero.  If  the  particular  instrument  will   display
overcorrection  as   a  negative  number,  this  spiking  procedure will  not  be
necessary.

      5.8  The quality control sample should be prepared in  the same acid matrix
as the calibration standards at 10 times the instrumental  detection limits and
in accordance with the instructions provided by the supplier.


6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See the material  in Chapter  Three,  Metallic   Analytes, Steps  3.1
through 3.3.


7.0  PROCEDURE

      7.1   Preliminary treatment  of most  matrices  is  necessary  because of the
complexity and variability of sample matrices.   Water  samples  which  have been
prefiltered and acidified will not need acid digestion as long as the samples and
standards  are matrix matched.   Solubilization  and digestion procedures  are
presented in Sample Preparation Methods (Methods 3005A-3050A).

      7.2  Set up the instrument with proper operating parameters established in
Step 4.2.  The instrument  must  be  allowed to  become thermally  stable before
beginning  (usually  requiring  at   least  30  minutes   of   operation  prior  to
calibration).

      7.3   Profile  and calibrate  the  instrument according  to  the  instrument
manufacturer's  recommended procedures,  using the  typical  mixed  calibration
standard solutions described in  Step 5.4.  Flush the  system with the calibration
blank (Step 5.5.1) between  each  standard or as  the manufacturer recommends. (Use
the average intensity of multiple  exposures for both standardization and sample
analysis to reduce  random  error.)   The calibration curve  should consist of a
blank and three standards.

      7.4   Before  beginning  the  sample run,  reanalyze  the  highest  mixed
calibration standard as if it were  a  sample. Concentration values obtained should
not deviate from the actual values by more than 5% (or the established control
limits, whichever  is  lower).  If  they do, follow  the recommendations  of  the
instrument manufacturer to correct for this condition.

      7.5   Flush the  system with  the calibration blank solution for  at least
1 minute (Step 5.5.1)  before  the analysis of each sample (see Note to Step 7.3).
Analyze the instrument check  standard (Step 5.6) and  the calibration blank (Step
5.5.1) after each 10 samples.
                                  6010A - 11                      Revision 1
                                                                  July 1992

-------
8.0  QUALITY CONTROL

      8.1  All  quality control  data  should be maintained and available for easy
reference or inspection.  Refer to  Chapter  One  for  additional  quality control
procedures.

      8.2   Dilute  and reanalyze samples that  are more concentrated  than the
linear calibration  limit  or use an  alternate,  less sensitive line  for which
quality control data is already established.

      8.3  Employ a  minimum  of one method blank per sample batch to determine if
contamination or any memory effects are occurring.  A method blank is a volume
of reagent water acidified with the  same  amounts of acids as were the standards
and samples.

      8.4   Analyze  one  replicate  sample  for every  twenty  samples  or  per
analytical batch, whichever is more frequent.   A  replicate sample is a sample
brought through the whole sample preparation  and analytical process  in duplicate.
Refer to Chapter One for a more detailed description of an  analytical batch.

      8.5   It  is recommended  that  whenever a new or  unusual  sample matrix is
encountered, a  series of tests be performed prior to reporting concentration data
for analyte elements.  These tests,  as  outlined  in Steps  8.5.1  and 8.5.2, will
ensure the analyst that neither positive nor negative  interferences  are operating
on any of the analyte elements to distort the accuracy of the reported values.

            8.5.1  Serial dilution:  If  the analyte  concentration is sufficiently
      high  (minimally,  a  factor of 10 above the  instrumental  detection limit
      after dilution), an analysis of  a 1:4 dilution should agree within + 10%
      of the original  determination. If not, a chemical  or physical interference
      effect should be suspected.

            8.5.2  Post digestion spike  addition: An  analyte  spike  added to a
      portion  of a  prepared sample, or  its dilution, should  be  recovered to
      within 75% to 125% of the known  value. The spike addition should produce
      a minimum level  of  10 times  and  a  maximum of  100 times the instrumental
      detection  limit. If  the spike  is not  recovered  within  the specified
      limits, a  matrix effect should be suspected.

            CAUTION:    If  spectral overlap is  suspected,  use of computerized
                        compensation,  an alternate  wavelength,  or  comparison
                        with an alternate method is  recommended.

      8.6  Check the instrument standardization by analyzing appropriate check
standards  as follows.

            8.6.1  Verify  calibration  every 10 samples and at the  end of the
      analytical  run,  using  a  calibration blank  (Step  5.5.1)  and  a check
      standard  (Step 5.6).

                  8.6.1.1  The results  of the check standard are to agree within
            10% of the expected value; if not, terminate the analysis,  correct
            the problem,  and reanalyze the  previous ten samples.

                                  6010A  - 12                      Revision 1
                                                                  July  1992

-------
                  8.6.1.2  The  results  of the calibration blank  are  to agree
            within three standard deviations of the mean  blank value.  If not,
            repeat the analysis two more times and average the results. If the
            average is not within three  standard  deviations  of the background
            mean, terminate the analysis,  correct the problem, recalibrate, and
            reanalyze the previous 10 samples.

            8.6.2   Verify the interelement and background  correction factors at
      the beginning and end of  an analytical run  or  twice during every 8-hour
      work  shift, whichever   is  more  frequent.   Do   this   by  analyzing  the
      interference check solution (Step  5.7). Results should  be within ± 20% of
      the true value obtained in Step 8.6.1.1.

            8.6.3   Spiked replicate samples are to be analyzed at a frequency
      of 5% or per analytical  batch, whichever is  more frequent.

                  8.6.3.1  The relative percent difference  between  replicate
            determinations is to be calculated as  follows:

                              Dl ' D2
                       RPD =
                             (D  + D
            where:

                  RPD = relative percent difference.
                   D1 = first sample value.
                   D2 = second sample value  (replicate).


                  (A control limit of + 20%  RPD  shall be used for sample values
            greater than ten times the instrument detection limit.)

                  8.6.3.2  The spiked replicate  sample  recovery is to be within
            + 20% of the actual value.
9.0  METHOD PERFORMANCE

      9.1   In an EPA round-robin Phase  1  study, seven laboratories applied the
ICP technique to  acid-distilled water matrices that had been  spiked with various
metal concentrates.  Table 4  lists the true values,  the mean reported values, and
the mean percent relative standard deviations.

      9.2   In a single laboratory evaluation, seven wastes were analyzed for 22
elements by  this method. The mean percent  relative standard  deviation  from
triplicate analyses  for all  elements  and wastes was 9 ±  2%.  The mean percent
recovery of spiked elements  for all wastes was 93 + 6%. Spike levels ranged from
100 ug/L to 100 mg/L. The wastes included sludges  and industrial wastewaters.
                                  6010A - 13                      Revision 1
                                                                  July 1992

-------
10.0  REFERENCES

1.    Hinge, R.K.; Peterson, V.J.; Fassel, V.A. Inductively Coupled Plasma-Atomic
Emission Spectroscopy;  Prominent Lines (final  report, March 1977 -February 1978);
EPA-600/4-79-017, Environmental Research Laboratory, Athens, GA, March 1979; Ames
Laboratory: Ames IA.


2.    Test  Methods:  Methods for  Organic Chemical Analysis  of Municipal  and
Industrial Wastewater;  U.S.  Environmental  Protection agency. Office of Research
and Development.  Environmental Monitoring and  Support Laboratory. ORD Publication
Offices of Center for Environmental  Research  Information: Cincinnati, OH, 1982;
EPA-600/4-82-057.

3.    Patel, B.K.; Raab, G.A.; et al. Report on a Single Laboratory Evaluation
of Inductively Coupled Optical Emission Method 6010; EPA Contract No. 68-03-3050,
December 1984.

4.    Sampling  and  Analysis Methods  for  Hazardous   Waste   Combustion;  U.S.
Environmental Protection Agency; Air and Energy Engineering Research Laboratory,
Office of Research and Development: Research Triangle Park, NC, 1986; Prepared
by Arthur D. Little, Inc.

5.    Bowmand, P.W.J.M. Line Coincidence Tables for Inductively Coupled Plasma
Atomic Emission Spectrometry, 2nd ed.; Pergamon: 1984.

6.    Rohrbough,  W.G.;  et  al.  Reagent  Chemicals,  American  Chemical  Society
Specifications, 7th ed.; American Chemical Society: Washington, DC, 1986.          ^^

7.    1985 Annual Book of ASTM Standards. Vol. 11.01; "Standard  Specification for     '
Reagent Water"; ASTM: Philadelphia, PA, 1985; D1193-77.
                                  6010A  -  14                      Revision  1
                                                                  July  1992

-------
                                   TABLE 4.
                       ICP PRECISION AND ACCURACY DATA8
         Sample No.  1
Sample No. 2
Sample No. 3
Ele-
ment
Be
Mn
V
As
Cr
Cu
Fe
Al
Cd
Co
Ni
Pb
Zn
Sec
Mean Re-
True ported Meaa
Value Value SD D
(ug/L) (ug/L) (%)
750
350
750
200
150
250
600
700
50
700
250
250
200
40
733
345
749
208
149
235
594
696
48
512
245
236
201
32
6.2
2.7
1.8
7.5
3.8
5.1
3.0
5.6
12
10
5.8
16
5.6
21.9
True
Value
(ug/L)
20
15
70
22
10
11
20
60
2.5
20
30
24
16
6
ported
Value
(ug/L)
20
15
69
19
10
11
19
62
2.9
20
28
30
19
8.5
Mean Re- Mean Re-
Mean. True ported Mean.
SD15 Value Value SD D
(%) (ug/L) (ug/L) (%)
9.8
6.7
2.9
23
18
40
15
33
16
4.1
11
32
45
42
180
100
170
60
50
70
180
160
14
120
60
80
80
10
176
99
169
63
50
67
178
161
13
108
55
80
82
8.5
5.2
3.3
1.1
17
3.3
7.9
6.0
13
16
21
14
14
9.4
8.3
     all  elements were analyzed by all  laboratories.
      standard deviation.
°Results  for Se are from two laboratories.
                                  6010A - 15
                           Revision 1
                           July 1992

-------
                                                 METHOD  6010A
                     INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION  SPECTROSCOPY
                       I. H.O
                     acidified,  \  No
                    pre-filtered?
  I*  sample
oils .greases
   waxes?
  Ii  sample
  queous or
   •olid?
  7.1  U..
Mathad 300S
 7.2  Sat up
and stabilize
 instrument
  7 1  Use
Hathod 3020
and Mathod
   7000
                                           7.1 U..
                                         Hathod 3010
 7 3  Profile
and calibrate
 ins trument
                                     7  4 Reanalyze
                                     highest mixed
                                     calibration
                                       standard
                                       7.5 Flush
                                      system and
                                       analyze
                                       sample
                                                                               7.5  Analyze
                                                                             check  standard
                                                                             and calibration
                                                                               blank after
                                                                             each 10 samples
                                         Adjust
                                      instrument per
                                       manufacturer
                                      recommendations
                                                                                                                  i
                                                                              7 6  Calculate
                                                                             concentrations
                                                                                  Stop
                                                     6010A  -  16
                                                     Revision  1
                                                     July  1992

-------
                                  METHOD 6020

                INDUCTIVELY COUPLED PLASMA - MASS SPECTROMETRY
1.0  SCOPE AND APPLICATION

      1.1  Inductively coupled plasma-mass spectrometry  (ICP-MS) is applicable
to the determination of sub-^g/L concentrations of a large number of elements  in
water  samples  and  in  waste  extracts  or  digests  [1,2].     When  dissolved
constituents are required, samples must be  filtered  and acid-preserved prior  to
analysis.  No digestion  is required prior to analysis for dissolved elements  in
water samples.  Acid digestion prior to filtration and analysis  is  required for
groundwater, aqueous samples, industrial  wastes,  soils, sludges,  sediments, and
other solid wastes for which total (acid-leachable) elements are required.

      1.2  ICP-MS has been applied to the determination  of over  60 elements  in
various matrices.  Analytes for which EPA has  demonstrated the acceptability  of
Method 6020 in a multi-laboratory study on solid wastes  are listed in Table  1.
Acceptability of the method for an element was based  upon the multi-laboratory
performance compared with that of either  furnace  atomic absorption  spectroscopy
or inductively coupled plasma-atomic emission  spectroscopy.   It  should be noted
that  the multi-laboratory  study  was conducted in  1986.    Multi-laboratory
performance data for the  listed elements  (and others) are  provided in Section  9.
Instrument detection limits,  sensitivities, and linear ranges will vary with the
matrices,  instrumentation,  and  operating  conditions.    In  relatively  simple
matrices, detection limits will generally be below 0.02/vg/L.

      1.3  If Method 6020 is used to determine any analyte not  listed in Table
1,  it  is the responsibility  of the  analyst  to  demonstrate the accuracy and
precision of  the Method  in the waste  to be analyzed.   The  analyst  is always
required to  monitor potential  sources  of  interferences  and  take appropriate
action to ensure data of known quality (see Section 8.4).

      1.4    Use  of  this  method  is  restricted  to  spectroscopists  who  are
knowledgeable in the recognition and in the correction of spectral, chemical, and
physical interferences in ICP-MS.

      1.5   An  appropriate  internal  standard  is  required  for each  analyte
determined by ICP-MS.  Recommended internal standards are 6Li,  45Sc, 89Y,  103Rh,
115In,  159Tb,  165Ho,  and 209Bi.   The  lithium  internal  standard  should  have   an
enriched abundance of 6Li, so that interference from lithium native to the sample
is minimized.   Other  elements  may need to be used  as internal  standards when
samples contain significant amounts of the recommended internal  standards.

2.0  SUMMARY OF METHOD

      2.1 Prior  to analysis,  samples  which require  total  ("acid-leachable")
values must be digested using appropriate sample preparation methods (such   as
Methods 3005 - 3051).


                                    6020-1                         Revision 0
                                                                  September 1994

-------
      2.2  Method 6020 describes the multi-elemental determination of analytes
by ICP-MS.  The method measures ions produced by a radio-frequency inductively
coupled plasma.  Analyte species originating in a liquid are nebulized and the
resulting aerosol  transported  by argon gas  into  the plasma torch.   The ions
produced  are  entrained  in the  plasma gas  and  introduced,  by  means  of  an
interface, into a mass spectrometer.  The ions produced in the plasma are sorted
according to their mass-to-charge ratios and  quantified with a channel electron
multiplier.   Interferences  must be assessed and valid corrections applied or the
data  flagged  to  indicate  problems.    Interference  correction  must  include
compensation for background ions  contributed by  the plasma gas,  reagents, and
constituents of the sample matrix.

3.0  INTERFERENCES

      3.1  Isobaric elemental  interferences  in  ICP-MS  are caused by isotopes of
different elements forming atomic ions with the same nominal mass-to-charge ratio
(m/z). A  data  system must be used  to  correct for  these  interferences.   This
involves determining the signal  for another  isotope of the interfering element
and subtracting the appropriate signal  from  the analyte isotope signal.  Since
commercial ICP-MS  instruments nominally provide  unit  resolution  at 10% of the
peak height, very high  ion currents at adjacent masses can also contribute to ion
signals at the mass of interest.  Although this type of interference  is uncommon,
it is not easily corrected, and samples exhibiting a significant problem of this
type could require resolution  improvement, matrix separation, or analysis using
another verified and documented isoptope,  or use of another method.

      3.2 Isobaric molecular and doubly-charged ion interferences in ICP-MS are
caused by ions consisting of more than one atom or charge, respectively.  Most
isobaric  interferences  that  could  affect  ICP-MS determinations have  been
identified in the  literature  [3,4].   Examples include ArCl+  ions  on the 75As
signal and  MoO+  ions  on the cadmium  isotopes.    While  the approach  used  to
correct for molecular isobaric  interferences is demonstrated  below  using the
natural isotope abundances from the literature [5], the most precise coefficients
for an instrument can  be determined  from  the ratio of the net  isotope signals
observed  for  a standard  solution at a  concentration providing suitable (<1
percent)  counting  statistics.    Because  the 35C1  natural  abundance  of  75.77
percent  is  3.13  times  the 37C1  abundance  of  24.23  percent,  the  chloride
correction for arsenic can be calculated (approximately) as follows (where the
38Ar37Cl+  contribution  at m/z  75  is a negligible 0.06 percent  of the  40Ar35Cl +
signal):

      corrected  arsenic   signal   (using   natural   isotopes   abundances  for
      coefficient approximations) =

      (m/z 75 signal)  - (3.13) (m/z 77 signal) + (2.73) (m/z 82 signal),
      (where the final term adjusts for any  selenium contribution at 77 m/z),

      NOTE:  Arsenic values can  be biased high by this  type of equation when the
      net signal  at m/z 82 is caused by ions other than   Se+, (e.g., 81BrH+ from
      bromine wastes [6]).


                                    6020-2                        Revision 0
                                                                  September 1994

-------
Similarly,

      corrected  cadmium   signal   (using  natural   isotopes   abundances  for
      coefficient approximations) =

      (m/z 114 signal) - (0.027)(m/z 118 signal) - (1.63)(m/z 108 signal),
      (where last 2 terms adjust for  any  tin or MoO+ contributions at m/z 114).

      NOTE: Cadmium  values will be  biased  low  by this type  of equation when
      92ZrO+  ions  contribute  at m/z 108,  but  use of  m/z  111 for  Cd  is even
      subject to direct  (94ZrOH+) and indirect  (90ZrO+) additive interferences
      when Zr is present.

      NOTE: As  for  the  arsenic equation  above,  the  coefficients  in  the  Cd
      equation are ONLY illustrative. The  most  appropriate coefficients for an
      instrument can  be  determined  from the ratio of  the  net isotope signals
      observed for a  standard  solution at a concentration providing suitable (<1
      percent) counting precision.

The accuracy  of  these types  of  equations  is  based upon the  constancy of the
OBSERVED isotopic ratios for the interfering species.   Corrections that presume
a constant fraction of a molecular ion relative to the "parent" ion have  not been
found [7] to be reliable, e.g., oxide levels can vary.  If a correction for an
oxide ion  is based  upon  the ratio  of  parent-to-oxide  ion  intensities,  the
correction must be adjusted for the degree of oxide formation by the use of an
appropriate oxide  internal  standard  previously  demonstrated  to form a similar
level  of oxide as the interferant. This type of correction has  been reported [7]
for oxide-ion corrections  using ThO+/Th+  for the  determination of  rare earth
elements.  The use of aerosol desolvation and/or mixed plasmas have been shown
to greatly reduce  molecular interferences  [8],   These techniques can  be used
provided that method  detection limits, accuracy, and precision requirements for
analysis of the samples can be met.

      3.3  Physical interferences are  associated with the sample nebulization and
transport processes as well  as with  ion-transmission efficiencies.  Nebulization
and transport processes  can be affected if a matrix component causes  a change in
surface  tension   or   viscosity.    Changes  in   matrix composition  can  cause
significant signal  suppression or enhancement [9].   Dissolved solids  can deposit
on the  nebulizer  tip of a  pneumatic nebulizer  and on the  interface  skimmers
(reducing the orifice size  and the instrument performance).  Total solid levels
below 0.2% (2,000 mg/L)  have been currently recommended [10]  to minimize solid
deposition.   An   internal  standard  can  be   used  to correct  for  physical
interferences, if it  is carefully matched to the analyte so that the two elements
are similarly affected by matrix changes [11].   When the  intensity level of an
internal standard  is  less  than 30  percent or greater  than 120  percent of the
intensity of  the  first  standard used during  calibration,  the  sample  must  be
reanalyzed after a fivefold (1+4) or greater dilution has been performed.

      3.4   Memory interferences can occur when there are  large concentration
differences between samples  or standards which are analyzed sequentially.  Sample


                                    6020-3                        Revision 0
                                                                  September 1994

-------
deposition on the sampler and skimmer cones,  spray  chamber design,  and the type
of nebulizer affect the extent of the memory interferences which are observed.
The rinse period  between  samples  must be long  enough to eliminate significant
memory interference.

4.0  APPARATUS AND MATERIALS

      4.1  Inductively coupled plasma-mass spectrometer:

            4.1.1   A system capable  of providing resolution,  better  than or
      equal to amu at 10% peak height is required.  The  system must have a mass
      range from at least  6  to 240 amu and a data system  that allows corrections
      for isobaric  interferences  and  the application of the internal standard
      technique.  Use of  a  mass-flow controller for  the nebulizer argon and a
      peristaltic pump for the sample solution are recommended.

            4.1.2  Argon gas supply:  high-purity grade (99.99%).

5.0  REAGENTS

      5.1  Acids used in  the preparation of standards  and for sample processing
must be of high purity.   Redistilled acids are recommended because of the high
sensitivity of ICP-MS.  Nitric acid at less than  2 per cent (v/v)  is  required for
ICP-MS to minimize damage to the interface and to minimize isobaric molecular-ion
interferences with  the  analytes.   Many more molecular-ion  interferences are
observed on the analytes  when  hydrochloric and  sulfuric acids are used [3,4].
Concentrations of antimony and silver between 50-500 /jg/l require  1% (v/v) HC1
for stability;  for concentrations  above  500 /jq/L  Ag,  additional  HC1  will  be
needed.

      5.2  Reagent water:  All references to water in the method refer to reagent
water unless  otherwise  specified.   Refer to Chapter One for  a definition of
reagent water.

      5.3  Standard stock  solutions may be purchased or prepared  from ultra-high
purity grade chemicals or metals (99.99 or greater  purity ).  See Method 6010A,
Section 5.3, for  instructions on preparing standard solutions from solids.

            5.3.1  Bismuth internal  standard solution, stock,  1  mL  =  100//g Bi:
      Dissolve 0.1115 g Bi203  in  a  minimum amount of dilute  HN03.   Add 10 ml
      cone. HN03 and dilute to 1,000 ml with reagent  water.

            5.3.2  Holmium internal  standard solution, stock,  1  mL  =  100//g Ho:
      Dissolve 0.1757 g Ho2(C03)2-5H20 in  10 mL  reagent water and 10 mL HN03.
      After dissolution  is  complete, warm the  solution to degas.   Add 10 mL
      cone. HN03 and dilute to 1,000 mL with reagent  water.

            5.3.3   Indium internal  standard  solution, stock, 1 mL  = 100 ^g In:
      Dissolve 0.1000 g indium  metal  in 10 mL  cone.  HN03.   Dilute to 1,000 mL
      with reagent water.


                                    6020-4                         Revision  0
                                                                   September 1994

-------
            5.3.4  Lithium internal standard solution,  stock,  1  ml = 100 fjg 6Li:
      Dissolve 0.6312 g 95-atom-% 6Li, Li2C03 in  10 ml of reagent water and 10 ml
      HN03.  After  dissolution is complete, warm the  solution  to degas.   Add
      10 ml cone. HN03 and dilute to 1,000 ml with reagent water.

            5.3.5  Rhodium internal standard solution, stock, 1 ml = 100 //g Rh:
      Dissolve 0.3593  g  ammonium hexachlororhodate (III)  (NH4)3RhCl6  in  10 ml
      reagent water.  Add 100  ml  cone. HC1  and  dilute  to  1,000  ml with reagent
      water.

            5.3.6  Scandium internal  standard solution, stock, 1 ml = 100/yg Sc:
      Dissolve 0.15343 g Sc203  in  10  ml  (1+1)  hot HN03.  Add 5 ml cone.  HN03 and
      dilute to 1,000 ml with  reagent water.

            5.3.7  Terbium internal standard solution, stock, 1 ml = 100//g Tb:
      Dissolve 0.1828 g Tb2(C03)3-5H20 in 10 ml (1+1) HN03.  After dissolution is
      complete, warm the solution  to degas.   Add 5 ml  cone.  HN03 and dilute to
      1,000 ml with reagent water.

            5.3.8  Yttrium internal  standard  solution, stock,  1 ml  =  100 /yg Y:
      Dissolve 0.2316 g Y2(C03)3.3H20 in  10 ml (1+1)  HN03.   Add  5 ml cone. HN03
      and dilute to 1,000 ml with  reagent  water.

            5.3.9  Titanium solution, stock, 1 ml = 100 fjg Ti:  Dissolve  0.4133 g
      (NH4)2TiF6 in  reagent water.  Add 2 drops  cone. HF  and dilute  to  1,000 ml
      with reagent water.

            5.3.10   Molybdenum solution,  stock,  1  mL =  100  jjg Mo:   Dissolve
      0.2043 g (NH4)2Mo04 in reagent water.  Dilute  to  1,000 mL with  reagent
      water.

      5.4  Mixed calibration  standard  solutions are prepared by diluting  the
stock-standard solutions to levels in the  linear range for the  instrument in a
solvent consisting of  1  percent (v/v)  HN03 in reagent water.   The  calibration
standard  solutions  must contain  a  suitable  concentration  of  an  appropriate
internal standard for each analyte.  Internal  standards may be added on-line at
the time  of analysis  using  a  second channel  of the  peristaltic  pump  and  an
appropriate mixing manifold.)   Generally, an internal standard should be no more
than 50 amu removed  from  the  analyte.   Recommended internal  standards include
6Li,  45Sc, 89Y,  103Rh, 1l5In, 159Tb,  ^Ho,  and 209Bi.   Prior  to preparing the mixed
standards, each stock solution  must be analyzed separately to determine  possible
spectral interferences or the  presence of  impurities.  Care must be taken when
preparing  the  mixed standards that  the elements  are compatible  and  stable.
Transfer the mixed standard solutions to freshly acid-cleaned FEP fluorocarbon
bottles for storage.  Fresh mixed standards must be prepared  as  needed  with the
realization that concentrations can change  on aging.   Calibration standards must
be initially verified  using  a  quality control  standard  (see Section  5.7)  and
monitored weekly for stability.
                                    6020-5                         Revision  0
                                                                   September 1994

-------
      5.5  Blanks:  Three  types of blanks are required  for  the  analysis.   The
calibration  blank  is  used  in  establishing   the  calibration  curve.    The
preparation blank is used to monitor for possible contamination resulting from
the sample preparation procedure.  The rinse blank is used to flush the system
between all samples and standards.

            5.5.1  The calibration blank consists of the same concentration(s)
      of the same acid(s)  used to prepare the final dilution of the calibrating
      solutions of  the  analytes [often  1 percent HN03  (v/v)  in  reagent water]
      along with  the  selected concentrations of  internal  standards  such  that
      there  is  an  appropriate  internal  standard  element  for  each of  the
      analytes.  Use of HC1 for antimony and silver is cited in Section 5.1

            5.5.2   The  preparation  (or  reagent)  blank  must be carried through
      the  complete preparation procedure  and   contain  the  same volumes  of
      reagents as the sample solutions.

            5.5.3   The  rinse blank consists of 1 to 2  percent  HN03  (v/v)  in
      reagent water.  Prepare a sufficient  quantity to flush the system between
      standards and samples.

            NOTE:  The  ICS  solutions  in  Table  2  are  intended  to  evaluate
            corrections for known interferences  on only the analytes in Table 1.
            If Method 6020 is used to determine an element not listed in Table
            1,  it  is  the responsibility  of the  analyst to  modify  the  ICS
            solutions, or prepare an alternative ICS solution,  to allow adequate
            verification of correction of interferences  on the unlisted element
            (see section 8.4).

      5.6  The  interference  check solution  (ICS)  is  prepared to contain known
concentrations of  interfering  elements  that will  demonstrate the magnitude of
interferences and provide an adequate test  of any corrections.  Chloride in the
ICS provides  a means  to evaluate  software corrections  for chloride-related
interferences such  as 35C11V  on 5V and  40Ar35Cl+  on  75As+.   Iron is used to
demonstrate adequate  resolution of the   spectrometer for the determination of
manganese. Molybdenum serves to indicate oxide effects on cadmium isotopes.  The
other components are present to evaluate the ability of the measurement system
to correct for various molecular-ion isobaric interferences.  The ICS is used to
verify that  the interference levels are corrected  by the data  system within
quality control limits.

            5.6.1   These  solutions  must be prepared  from ultra-pure  reagents.
      They can be obtained commercially  or  prepared by the following procedure.

                  5.6.1.1   Mixed  ICS  solution  I may  be  prepared by adding
            13.903 g A1(N03)3-9H20,  2.498 g CaC03 (dried at 180 C for 1 h before
            weighing), 1.000 g Fe, 1.658 g MgO, 2.305 g Na2C03, and 1.767 g K2C03
            to 25 ml of reagent water.    Slowly  add 40 mL of  (1+1) HN03.  After
            dissolution  is  complete,  warm  the  solution  to  degas.    Cool  and
            dilute to 1,000 ml  with reagent water.


                                    6020-6                        Revision  0
                                                                  September  1994

-------
                  5.6.1.2   Mixed  ICS  solution  II  may be  prepared by  slowly
            adding 7.444 g 85 % H3P04,  6.373 g 96% H2S04, 40.024 g 37%  HC1,  and
            10.664 g citric  acid  C607H8 to 100 ml of reagent water.  Dilute to
            1,000 ml with reagent water.

                  5.6.1.3   Mixed ICS  solution  III may be  prepared by  adding
            1.00 ml  each  of  100-/yg/mL  arsenic,  cadmium,   chromium,  cobalt,
            copper, manganese, nickel,  silver, and zinc stock solutions to about
            50 ml  reagent  water. Add  2.0  ml concentrated  HN03,  and dilute  to
            100.0 ml with reagent water.

                  5.6.1.4  Working  ICS Solutions

                        5.6.1.4.1   ICS-A  may be prepared  by adding 10.0 ml  of
                  mixed  ICS solution  I   (5.7.1.1),  2.0 ml  each  of  100-/vg/mL
                  titanium stock solution (5.3.9) and molybdenum stock  solution
                  (5.3.10),  and   5.0 ml  of  mixed  ICS  solution  II  (5.7.1.2).
                  Dilute to  100 ml  with reagent water.  ICS  solution A must be
                  prepared fresh weekly.

                        5.6.1.4.2   ICS-AB may be prepared by adding 10.0 ml of
                  mixed  ICS solution  I   (5.7.1.1),  2.0 ml  each  of  100-//g/mL
                  titanium stock solution (5.3.9) and molybdenum stock  solution
                  (5.3.10),  5.0  ml of  mixed ICS  solution   II  (5.7.1.2),  and
                  2.0 ml of Mixed ICS solution III  (5.7.1.3).   Dilute to  100 ml
                  with  reagent water.   Although  the ICS solution  AB must  be
                  prepared fresh  weekly,  the analyst should be aware  that  the
                  solution may precipitate silver more quickly.

      5.7  The quality control  standard is the initial calibration verification
solution (ICV), which must be prepared in  the  same acid matrix as the  calibration
standards.  This solution must be an independent standard near the midpoint of
the  linear range  at  a concentration  other  than  that  used for  instrument
calibration.  An independent standard  is defined as  a standard composed  of  the
analytes from a source  different from those used in the  standards for instrument
calibration.

      5.8  Mass  spectrometer tuning solution.   A solution containing  elements
representing all  of the mass regions of interest  (for example, 10/yg/L of Li,  Co,
In, and Tl) must  be prepared to verify  that the  resolution and mass  calibration
of the instrument are within the required specifications (see Section 7.5). This
solution is also  used to verify that  the instrument has reached thermal stability
(See Section 7.4).

6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Sample  collection procedures should  address  the  considerations
described in Chapter Nine of this Manual.
                                    6020-7                        Revision  0
                                                                  September 1994

-------
      6.2  See the introductory material in Chapter Three, Inorganic Analytes,
Sections  3.1.3  for  information  on  sample  handling  and  preservation.   Only
polyethylene or fluorocarbon (TFE or PFA) containers are recommended for use in
Method 6020.

7.0   PROCEDURE

      7.1  Solubilization and digestion procedures are presented in the Sample
Preparation Methods (e.g., Methods 3005 - 3051).

      7.2   Initiate  appropriate operating configuration of  the  instruments
computer according to the instrument manufacturer's instructions.

      7.3  Set up the instrument  with the proper operating parameters according
to the instrument manufacturer's instructions.

      7.4   Operating  conditions:  The analyst  should  follow the instructions
provided by the  instrument manufacturer.   Allow  at  least 30  minutes  for the
instrument to equilibrate before analyzing any samples.  This must be verified
by analyzing a tuning solution (Section 5.8) at least four times with relative
standard deviations of < 5% for the analytes contained in the tuning solution.

            NOTE:   Precautions must be taken to  protect  the channel  electron
            multiplier from high  ion currents.   The channel electron multiplier
            suffers from fatigue  after  being exposed to high  ion currents.  This
            fatigue  can  last  from  several  seconds  to hours depending  on the
            extent of exposure.  During this time period, response factors are
            constantly changing,  which  invalidates the calibration curve, causes
            instability, and invalidates sample analyses.

      7.5  Conduct mass calibration  and resolution checks in  the mass regions of
interest.  The mass calibration and  resolution parameters  are required criteria
which must be met prior to any  samples  being analyzed.   If the mass calibration
differs more than 0.1  amu  from the true value,  then the mass calibration must be
adjusted to the correct value.  The  resolution must also be verified to be less
than 0.9 amu full width at 10 percent peak height.

      7.6   Calibrate  the  instrument for the analytes  of  interest (recommended
isotopes  for  the  analytes  in Table  1 are  provided   in  Table 3),  using the
calibration blank and at least a single initial  calibration standard according
to the instrument manufacturer's procedure.  Flush  the system with the rinse
blank (5.5.3)  between each standard  solution.  Use the  average  of at leastthree
integrations for both calibration and sample analyses.

      7.7   All masses which could  affect data  quality should  be monitored to
determine potential effects from matrix  components on the analyte peaks.   The
recommended isotopes to be monitored are liste in Table 3.

      7.8    Immediately  after  the calibration  has  been  established,  the
calibration must be verified and  documented for every analyte by the analysis of
the calibration verification solution  (Section 5.7).   When measurements exceed

                                    6020-8                        Revision  0
                                                                  September 1994

-------
±  10% of the  accepted value,  the analyses must  be terminated,  the problem
corrected, the instrument recalibrated, and the new calibration verified.  Any
samples analyzed under an out-of-control calibration must be reanalyzed.  During
the course of an analytical run, the instrument may be "resloped"  or recalibrated
to  correct  for  instrument  drift.   A  recalibration must  then  be followed
immediately by a new analysis of a  CCV and CCB before any further samples may be
analyzed.

      7.9   Flush  the system  with  the  rinse blank  solution  (5.5.3)  until the
signal levels return  to the  method's  levels of quantitation (usually about 30
seconds) before the analysis  of  each sample (see Section 7.7).   Nebulize each
sample until a steady-state signal  is  achieved  (usually  about 30 seconds) prior
to collecting data.  Analyze the calibration verification solution (Section 5.6)
and the calibration blank (Section  5.5.1)  at a  frequency of at least  once every
10 analytical samples.  Flow-injection systems may be used as long as they can
meet the performance criteria of this method.

      7.10   Dilute  and reanalyze samples that are  more concentrated than the
linear range for an analyte  (or  species needed for a correction) or measure an
alternate less-abundant isotope.   The  linearity  at  the  alternate mass must be
confirmed by appropriate calibration (see Sec.  7.6 and 7.8).

      7.11   Calculations:    The  quantitative values  shall  be  reported  in
appropriate units, such as micrograms per liter (jjg/L) for aqueous samples and
milligrams per kilogram (mg/kg)  for solid samples.   If dilutions were performed,
the appropriate corrections must be applied to the sample values.

            7.11.1  If appropriate, or required, calculate results for solids on
      a dry-weight basis as follows:

                  (1)   A  separate determination  of percent  solids must  be
                        performed.

                  (2)   The concentrations  determined in  the  digest  are to be
                        reported on the basis of the dry weight of the sample.

                  Concentration  (dry weight)(mg/kg) =  •: x
                                                       W X o

                        Where,

                  C = Digest Concentration  (mg/L)
                  V = Final volume  in liters after sample preparation
                  W = Weight  in kg  of wet sample

                       % Solids
                          100

      Calculations  should  include   appropriate  interference corrections  (see
      Section  3.2  for  examples),   internal-standard normalization,  and  the


                                    6020-9                        Revision 0
                                                                  September 1994

-------
      summation of signals at 206,  207,  and  208 m/z for lead (to compensate for
      any differences  in  the  abundances  of  these  isotopes between samples and
      standards).

8.0  QUALITY CONTROL

      8.1  All quality control  data  should  be maintained and  be available for
easy reference or inspection.

      8.2   Instrument  Detection  Limits  (IDLs)  in /jg/l  can  be  estimated  by
calculating the average of the  standard  deviations of the three runs on three
non-consecutive days from the analysis  of a reagent  blank solution with seven
consecutive measurements per day.   Each measurement must  be performed as though
it were a separate analytical sample (i.e., each measurement must be followed by
a rinse  and/or any other  procedure normally performed between the analysis of
separate samples).  IDLs must be determined at least every three  months and kept
with the instrument log book.  Refer to Chapter One for additional guidance.

      8.3  The intensities of all internal standards must  be monitored for every
analysis.  When the  intensity of any  internal  standard fails to  fall between 30
and  120  percent  of  the  intensity of  that   internal  standard  in  the  initial
calibration standard,  the following procedure is  followed.  The sample must be
diluted fivefold  (1+4) and reanalyzed with the addition of appropriate amounts
of  internal  standards.  This procedure  must be  repeated  until  the  internal -
standard intensities  fall  within the prescribed window. The intensity levels of
the internal  standards for the calibration blank (Section  5.5.1) and instrument
check standard  (Section 5.6)  must  agree within ± 20  percent  of the  intensity
level of the  internal standard of the original calibration solution.  If they do
not agree, terminate  the analysis,  correct the problem, recalibrate, verify the
new calibration,  and reanalyze the affected  samples.

      8.4  To obtain analyte data of known quality, it is necessary to measure
more than the analytes  of  interest  in order to apply corrections  or to determine
whether  interference  corrections  are  necessary.    If  the concentrations  of
interference sources  (such as  C, Cl, Mo, Zr,  W) are  such that,  at the correction
factor,  the  analyte  is   less  than  the  limit  of  quantification  and  the
concentration  of  interferents  are  insignificant,   then  the  data  may  go
uncorrected.   Note that monitoring the interference  sources does  not necessarily
require monitoring the interferant itself, but that a molecular species may be
monitored  to indicate the  presence  of  the interferent.    When  correcttion
equations  are used,   all  QC  criteria  must  also  be  met.   Extensive  QC  for
interference corrections are required at all times.  The monitored masses must
include  those elements whose hydrogen,  oxygen, hydroxyl,  chlorine,  nitrogen,
carbon  and  sulfur  molecular  ions  could  impact  the  analytes of  interest.
Unsuspected interferences  may be detected by adding  pure major matrix components
to a sample to observe any impact on  the  analyte signals.  When  an interference
source is present, the sample elements impacted must  be flagged  to indicate (a)
the percentage interference correction applied to the  data or (b) an uncorrected
interference  by virtue of the elemental equation used for quantitation.   The
isotope proportions  for an element  or molecular-ion cluster provide information
useful for quality assurance.

                                   6020-10                        Revision  0
                                                                  September 1994

-------
      NOTE:  Only isobaric elemental, molecular, and doubly charged interference
      corrections which use the observed  isotopic-response ratios or parent-to-
      oxide ratios (provided an oxide internal  standard  is used as described in
      Section 3.2) for each  instrument system are acceptable  corrections for use
      in Method 6020.

      8.5  Dilution  Test:   If the analyte concentration  is within  the linear
dynamic range of the  instrument and sufficiently high (minimally,  a factor of at
least 100 times greater  than  the  concentration in the reagent blank,  refer to
Section 5.5.2), an analysis  of a fivefold (1+4) dilution  must agree within ± 10%
of the original  determination.  If not, an interference effect must be suspected.
One dilution  test  must be  included for each twenty  samples (or  less)  of each
matrix in a batch.

      8.6  Post-Digestion Spike Addition:  An analyte spike added to a portion
of a prepared sample, or its dilution, should be recovered to within 75 to 125
percent of the known  value or within the laboratory derived acceptance criteria.
The spike  addition  should  be based on  the  indigenous  concentration  of  each
element of interest  in the  sample.   If the  spike  is  not  recovered  within the
specified limits,  the sample must be diluted and reanalyzed to compensate for the
matrix effect.  Results must agree to within  10% of the original determination.
The use of a standard-addition analysis procedure  may also be used to compensate
for this effect (Refer to Method 7000).

      8.7  A  Laboratory Control Sample (LCS)  should be analyzed for each analyte
using the  same sample preparations,  analytical  methods and  QA/QC  procedures
employed for the test samples. One LCS should be prepared and analyzed for each
sample batch at a frequency of one LCS for each 20 samples or less.

      8.8  Check  the instrument calibration by  analyzing  appropriate  quality
control solutions as follows:

            8.8.1  Check  instrument   calibration   using  a  calibration  blank
      (Section  5.5.1)  and  the  initial  calibration  verification  solution
      (Sections 5.7 and 7.9).

            8.8.2  Verify  calibration at a  frequency of every  10  analytical
      samples  with  the  instrument  check   standard  (Section  5.6)  and  the
      calibration blank (Section 5.5.1).  These solutions must also be analyzed
      for each  analyte at  the  beginning of the  analysis  and after the  last
      sample.

            8.8.3  The results of  the initial calibration verification solution
      and the instrument check standard must  agree within ±  10% of the expected
      value.     If not,  terminate  the   analysis,   correct  the  problem,   and
      recalibrate the instrument.   Any sample analyzed under an out-of-control
      calibration must be reanalyzed .

            8.8.4  The  results of the calibration  blank  must be less than  3
      times the current  IDL for each element.   If this is not the case,  the
      reason  for the out-of-control  condition must be found and corrected,  and

                                   6020-11                        Revision  0
                                                                  September 1994

-------
      affected samples must be reanalyzed.  If the laboratory consistently has
      concentrations greater than 3 times the  IDL, the  IDL may be indicative of
      an estimated IDL and should be re-evaluated.

      8.9    Verify the  magnitude  of  elemental  and  molecular-ion  isobaric
interferences  and  the  adequacy  of  any  corrections  at the  beginning  of  an
analytical run or once every 12 hours, whichever is more frequent.  Do this by
analyzing the interference check solutions A and AB.  The analyst  should be aware
that precipitation from solution AB may occur with some elements, specifically
silver. Refer to Section 3.0 for  a discussion on  intereferences and potential
solutions to those intereferences if additional guidance is needed.
                                                                          at a
      8.10   Analyze one  duplicate sample  for  every matrix  in  a batch
frequency of one matrix duplicate for every 20 samples.

            8.10.1   The relative  percent  difference  (RPD)  between  duplicate
      determinations must be calculated as follows:

                                ID, - D2 |
                   RPD =      	     x 100
                               (D, + D2)/2

            where:

            RPD = relative percent difference.
            D, = first sample value.
            D2 = second sample value (duplicate)

      A control  limit of 20%  RPD should  not  be exceeded  for  analyte values
      greater than 100 times  the instrumental detection limit.  If this  limit is
      exceeded, the reason for the  out-of-control  situation must be found and
      corrected, and  any  samples  analyzed  during the out-of-control  condition
      must be reanalyzed.

9.0  METHOD PERFORMANCE

      9.1  In an EPA multi-laboratory study, 10 laboratories applied the
ICP-MS technique  to  both aqueous  and  solid samples.  TABLE  4  summarizes the
method performance data for aqueous samples.  Performance data for solid samples
is provided in TABLE 5.

10.0  REFERENCES

1. Horlick, G., et al., Spectrochim. Acta  40B, 1555  (1985).

2. Gray, A.L., Spectrochim.  Acta  40B, 1525  (1985); 41B,  151  (1986).

3. Tan, S.H., and Horlick, G., Appl. Spectrosc. 40,  445  (1986).

4. Vaughan, M.A., and  Horlick, G., Appl. Spectrosc.  40,  434  (1986).


                                    6020-12                       Revision  0
                                                                  September 1994

-------
5. Holden, N.E.,  "Table  of the Isotopes,"  in Lide, D.R., Ed., CRC Handbook of
Chemistry and Physics, 74th Ed., CRC Press, Boca Raton,  FL, 1993.
6. Hinners, T.A., Heithmar, E., Rissmann, E., and  Smith, D., Winter Conference
on Plasma Spectrochemistry, Abstract THP18; p. 237, San  Diego, CA  (1994).
7. Lichte, F.E., et al., Anal. Chem. 59, 1150 (1987).
8. Evans E.H., and Ebdon,  L., J. Anal. At. Spectrom. 4,  299 (1989).
9. Beauchemin, D., et al., Spectrochim. Acta 42B,  467  (1987).
10. Houk, R.S., Anal. Chem. 58, 97A (1986).
11. Thompson, J.J., and Houk, R.S., Appl. Spectrosc. 41, 801 (1987).
                                   6020-13                        Revision  0
                                                                  Septenter 1994

-------
TABLE 1. ELEMENTS APPROVED FOR  ICP-MS DETERMINATION
            Element                      CAS* #
            Aluminum                   7429-90-5
            Antimony                   7440-36-0
            Arsenic                    7440-38-2
            Barium                     7440-39-3
            Beryllium                  7440-41-7
            Cadmium                    7440-43-9
            Chromium                   7440-47-3
            Cobalt                     7440-48-4
            Copper                     7440-50-8
            Lead                       7439-92-1
            Manganese                  7439-96-5
            Nickel                     7440-02-0
            Silver                     7440-22-4
            Thallium                   7440-28-0
            Zinc                       7440-66-6
                                    6020-14                        Revision 0
                                                                   September 1994

-------
TABLE 2.  RECOMMENDED  INTERFERENCE  CHECK SAMPLE COMPONENTS AND CONCENTRATIONS
Solution
component
Al
Ca
Fe
Mg
Na
P
K
S
C
Cl
Mo
Ti
As
Cd
Cr
Co
Cu
Mn
Ni
Ag
Zn
Solution A
Concentration (mg/L)
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
200.0
1000.0
2.0
2.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Solution AB
Concentration (mg/L)
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
200.0
1000.0
2.0
2.0
0.0200
0.0200
0.0200
0.0200
0.0200
0.0200
0.0200
0.0200
0.0200
                                    6020-15                        Revision  0
                                                                   September 1994

-------
TABLE 3.  RECOMMENDED ISOTOPES FOR SELECTED ELEMENTS
Mass                                           Element of interest


27                                                   Aluminum
121, 123                                             Antimony
75                                                   Arsenic
138, 137, 136, 135. 134                              Barium
9                                                    Beryllium
209                                                  Bismuth (IS)
114. 112, 111. 110, 113, 116, 106                    Cadmium
42, 43, 44, 46, 48                                   Calcium (I)
35, 37, (77, 82)a                                    Chlorine (I)
52, 53, 50, 54                                       Chromium
59                                                   Cobalt
63, 65                                               Copper
165                                                  Holmium (IS)
115, 113                                             Indium (IS)
56, 54, 57, 58                                       Iron (I)
139                                                  Lanthanum (I)
208, 207. 206, 204                                   Lead
6*7 7                                                Lithium (IS)
24, 25, 26                                           Magnesium (I)
55                                                   Manganese
98, 96, 92, 97, 94, (108)a                           Molybdenum (I)
58, 60, 62, 61, 64                                   Nickel
39                                                   Potassium (I)
103                                                  Rhodium (IS)
45                                                   Scandium (IS)
107. 109                                             Silver
23                                                   Sodium (I)
159                                                  Terbium (IS)
205, 203                                             Thallium
120, 118                                             Tin (I)
89                                                   Yttrium (IS)
64, 66, 68, 67, 70                                   Zinc

       NOTE:  Method 6020 is recommended for only  those  analytes  listed  in Table
1.   Other  elements  are  included  in  this table  because  they  are potential
interferents (labeled  I) in the determination of recommended analytes, or because
they are commonly used internal  standards (labeled  IS).   Isotopes  are listed in
descending order of natural abundance.  The most generally useful isotopes are
underlined and  in  boldface,  although certain matrices may  require  the use of
alternative isotopes.  a These masses are also useful  for interference correction
(Section 3.2).  b Internal standard must be enriched in  the 6Li isotope.  This
minimizes interference from indigenous lithium.
                                    6020-16                       Revision  0
                                                                  September 1994
                                                                                   i

-------
TABLE  4.
SOLUTIONS
ICP-MS MULTI-LABORATORY  PRECISION AND  ACCURACY DATA  FOR AQUEOUS
Element
Comparability3
Range
%RSD
Range
Nb Sc
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
        95  -  100
           d
        97  -  114
        91  -  99
       103  -  107
        98  -  102
        99  -  107
        95  -  105
       101  -  104
        85  -  101
        91  -  900
        71  -  137
        98  -  102
        95  -  101
        98  -  101
       101  -  114
       102  -  107
       104  -  105
        82  -  104
        88  -  97
       107  -  142
        93  -  102
11
5.0
7.1
4.3
8.6
4.6
5.7
13
8.2
6.1
11
11
10
8.8
6.1
9.9
15
5.2
24
9.7
23
6.8
- 14
- 7.6
- 48
- 9.0
- 14
- 7.2
- 23
- 27
- 8.5
- 27
- 150
- 23
- 15
- 15
- 6.7
- 19
- 25
- 7.7
- 43
- 12
- 68
- 17
14 -
16 -
12 -
16 -
13 -
18 -
17 -
16 -
18 -
17 -
10 -
17 -
16 -
18 -
18 -
11 -
12 -
13 -
9 -
18 -
8 -
16 -
14
16
14
16
14
20
18
18
18
18
12
18
16
18
18
12
12
16
10
18
13
18
4
3
4
5
3
3
5
4
3
5
5
6
5
4
2
5
3
2
5
3
3
5
a Comparability refers to the percent agreement of mean ICP-MS values to those
of  the reference  technique,      N  is  the  range of  the  number   of  ICP-MS
measurements where the analyte values exceed the limit of quantitation  (3.3 times
the average IDL value).      c S is the number of samples with results greater
than the  limit of quantitation.   d No comparability values  are  provided  for
antimony  because  of  evidence  that  the  reference  data  is  affected  by  an
interference.
                                    6020-17
                                                      Revision  0
                                                      September 1994

-------
TABLE 5.   ICP-MS MULT I-LABORATORY PRECISION AND ACCURACY DATA  FOR SOLID MATRICES
Element
Comparability8
Range
%RSD
Range Nb
Sc
Aluminum
Antimony
Arsenic
Barium
Beryl 1i urn
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Nickel
Potassium
Selenium
Silver
Sodium
Thallium
Vanadium
Zinc
 83 - 101
    d
 79 - 102
100 - 102
 50 - 87
 93 - 100
 95 - 109
 77 - 98
 43 - 102
 90 - 109
 87 - 99
 90 - 104
 89 - 111
 80 - 108
 87 - 117
 97 - 137
   81
 43 - 112
100 - 146
   91
 83 - 147
 84 - 124
 11 - 39
 12 - 21
 12 - 23
4.3 - 17
 19 - 34
6.2 - 25
4.1 - 27
 11 - 32
 15 - 30
9.0 - 25
6.7 - 21
5.9 - 28
7.6 - 37
 11 - 40
9.2 - 29
 11 - 62
   39
 12 - 33
 14 - 77
   33
 20 - 70
 14 - 42
13 - 14
15 - 16
16 - 16
15 - 16
12 - 14
19 - 20
15 - 17
17 - 18
17 - 18
18 - 18
12 - 12
15 - 18
15 - 16
16 - 18
16 - 18
10 - 12
  12
15 - 15
 8 - 10
  18
 6 - 14
18 - 18
7
2
7
7
5
5
7
7
6
7
7
7
7
7
7
5
1
3
5
1
7
7
a Comparability refers to the percent agreement of mean ICP-MS values to those
of  the reference  technique.   b  N  is  the  range  of  the  number  of  ICP-MS
measurements where the analyte values exceed the limit of quantitation  (3.3 times
the average IDL value).      c S is the number of samples with results greater
than the  limit of quantitation.   d No comparability values  are  provided for
antimony  because  of evidence  that  the  reference  data  is  affected  by  an
interference.
                                    6020-18
                                                Revision 0
                                                September 1994

-------
                  METHOD 6020
INDUCTIVELY COUPLED PLASMA  - MASS  SPECTROMETRY
7.9 Anah/ia
cheek atandard
•nd calibration
blank afttr
•aeh 1 0 Mmpiaa.
4

7.9 Fluah
ayatom and
analyza
aampla

                   6020-19
Revision  0
September 1994

-------
                                 METHOD 7000A

                           ATOMIC  ABSORPTION  METHODS
1.0   SCOPE AND APPLICATION

      1.1   Metals in solution may  be  readily  determined  by atomic absorption
spectroscopy. The method is simple, rapid, and applicable to a large number of
metals  in  drinking,  surface,  and  saline waters  and domestic  and industrial
wastes. While drinking water free of particulate matter may  be analyzed directly,
ground  water,  other  aqueous samples,  EP extracts,  industrial  wastes,  soils,
sludges, sediments, and other solid wastes require digestion prior to analysis
for both total and acid  Teachable metals.  Analysis for dissolved elements does
not require digestion if the sample has been filtered and acidified.

      1.2   Detection limits,  sensitivity, and optimum ranges of the metals will
vary with the matrices and models of atomic absorption spectrophotometers. The
data shown in Table 1 provide some indication of the detection limits obtainable
by direct aspiration and by furnace techniques. For clean aqueous samples, the
detection limits shown in the table  by direct aspiration may be extended downward
with  scale  expansion and  upward  by using  a  less sensitive wavelength  or by
rotating the  burner  head.  Detection limits  by direct aspiration  may  also be
extended through concentration of the sample and/or through solvent extraction
techniques. For  certain  samples,  lower concentrations may  also be determined
using the furnace techniques. The detection limits given in Table 1  are somewhat
dependent  on  equipment  (such  as  the  type  of spectrophotometer  and  furnace
accessory, the energy source, the degree of electrical expansion of the output
signal), and are greatly dependent on sample matrix.   Detection  limits should be
established,  empirically,  for  each matrix type analyzed.   When using  furnace
techniques, however, the  analyst should be cautioned  as  to possible chemical
reactions  occurring   at  elevated  temperatures  which  may  result in  either
suppression or enhancement of  the  analysis  element.  To ensure  valid data with
furnace techniques, the analyst must examine each matrix for interference effects
(see  Step  3.2.1)  and,  if detected,  treat  them accordingly,  using  either
successive dilution,  matrix modification, or method of standard additions (see
Step 8.7).

      1.3   Where direct-aspiration atomic absorption techniques do not provide
adequate sensitivity, reference is made to specialized procedures  (in addition
to the  furnace  procedure)  such as the gaseous-hydride method  for  arsenic and
selenium and the cold-vapor technique for mercury.


2.0   SUMMARY OF METHOD

      2.1   Although methods have  been reported  for  the  analysis of solids by
atomic absorption spectroscopy, the technique generally is limited  to metals in
solution or solubilized through some form of sample processing.

      2.2   Preliminary treatment of waste water,  ground water,  EP extracts, and
industrial waste is always necessary because of the complexity and variability
of sample matrix. Solids,  slurries,  and suspended material must be subjected to
a solubilization process before analysis. This process may vary because of the

                                   7000A  -  1                       Revision 1
                                                                  July 1992

-------
metals to be determined and the nature  of the sample being analyzed.  Solubili-
zation and digestion procedures are  presented  in  Step  3.2  (Sample Preparation
Methods).

      2.3   In direct-aspiration  atomic  absorption  spectroscopy,  a  sample  is
aspirated and atomized  in  a flame. A light beam from  a hollow cathode lamp or an
electrodeless discharge lamp is directed through the  flame into a monochromator,
and onto  a detector that  measures the amount  of absorbed light.  Absorption
depends upon  the  presence  of free unexcited ground-state atoms  in the flame.
Because the wavelength  of  the light beam is characteristic of  only  the metal
being determined, the  light energy absorbed by the  flame is a  measure of the
concentration of that metal in the sample.  This principle is the basis of atomic
absorption spectroscopy.

      2.4   When  using  the  furnace  technique in conjunction  with  an atomic
absorption spectrophotometer, a representative aliquot  of a sample  is placed in
the graphite tube in the furnace,  evaporated to dryness, charred, and atomized.
As a greater percentage of available  analyte atoms is vaporized and dissociated
for absorption  in the  tube  rather than the flame,  the use of  smaller sample
volumes  or detection  of   lower concentrations  of  elements  is  possible.  The
principle  is essentially the same as with direct aspiration atomic absorption,
except that  a furnace, rather  than  a flame,  is  used  to atomize  the  sample.
Radiation  from a  given  excited  element is  passed  through  the  vapor containing
ground-state atoms of that  element. The intensity of the transmitted radiation
decreases  in proportion to the amount of the ground-state element in the vapor.
The metal atoms to be measured are placed in the beam of radiation by increasing
the temperature  of  the furnace,  thereby  causing  the  injected  specimen  to  be
volatilized.  A monochromator  isolates the  characteristic radiation from the
hollow cathode lamp  or electrodeless discharge lamp,  and a photosensitive device
measures the attenuated transmitted radiation.
3.0   INTERFERENCES

      3.1   Direct aspiration

            3.1.1   The  most  troublesome  type  of  interference  in  atomic
      absorption spectrophotometry is usually termed "chemical" and is caused by
      lack of absorption of atoms bound in molecular combination in the flame.
      This  phenomenon  can  occur  when  the  flame is  not sufficiently  hot  to
      dissociate the molecule,  as in the case  of  phosphate interference with
      magnesium,  or  when  the  dissociated atom  is  immediately oxidized  to a
      compound that will  not dissociate  further at the temperature  of the flame.
      The  addition  of  lanthanum  will  overcome  phosphate   interference  in
      magnesium,   calcium,   and  barium  determinations.   Similarly,   silica
      interference in  the  determination of  manganese can be eliminated by the
      addition of calcium.

            3.1.2   Chemical interferences may  also be eliminated by separating
      the metal  from the interfering material.  Although complexing agents are
      employed primarily to increase  the  sensitivity of the analysis, they may
      also be used to  eliminate or reduce interferences.
                                   7000A - 2                      Revision 1
                                                                  July 1992

-------
      3.1.3    The presence of  high  dissolved  solids in the  sample  may
result  in  an   interference  from  nonatomic absorbance  such  as  light
scattering.  If  background correction  is  not available, a  nonabsorbing
wavelength should be checked.  Preferably,  samples containing high solids
should be extracted.

      3.1.4  lonization interferences occur when the  flame temperature is
sufficiently high to generate the  removal  of an electron from a neutral
atom, giving  a positively  charged ion. This  type  of  interference  can
generally  be  controlled  by the addition,   to  both  standard  and sample
solutions, of a large  excess  (1,000 mg/L)   of an  easily  ionized element
such as K, Na,  Li or Cs.

      3.1.5  Spectral  interference can occur when an  absorbing wavelength
of an element present in the sample but  not  being determined falls within
the width of the absorption line of the element of interest. The results
of  the  determination  will  then  be  erroneously  high,   due  to  the
contribution of the interfering element to the  atomic absorption signal.
Interference can also occur when resonant energy from another element in
a multielement lamp, or from a metal impurity in the lamp cathode, falls
within the bandpass of the slit setting when that other metal  is present
in the  sample.  This type of  interference  may  sometimes be  reduced by
narrowing the slit width.

      3.1.6  Samples  and  standards  should  be  monitored for  viscosity
differences that may alter the aspiration rate.

      3.1.7  All  metals  are  not  equally  stable  in  the  digestate,
especially  if  it  contains  only  nitric   acid,   not  nitric  acid  and
hydrochloric acid.  The digestate  should be analyzed  as soon as possible,
with preference given to  Sn, Sb, Mo, Ba, and Ag.

3.2   Furnace procedure

      3.2.1   Although the problem of oxide formation is greatly reduced
with furnace procedures because atomization occurs in  an inert atmosphere,
the technique is still  subject to  chemical  interferences. The composition
of the sample matrix can have  a  major effect on  the analysis. It  is those
effects  which  must be  determined and  taken  into consideration  in  the
analysis of each different matrix  encountered. To  help verify the absence
of matrix  or  chemical  interference,  the serial  dilution technique  (see
Step  8.6)  may  be  used.  Those  samples which  indicate  the presence of
interference should be treated in one or more of the following ways:

      1.    Successively  dilute  and  reanalyze  the samples  to eliminate
            interferences.

      2.    Modify the sample  matrix  either to remove interferences or to
            stabilize the analyte. Examples are the  addition of ammonium
            nitrate  to remove  alkali  chlorides  and  the  addition  of
            ammonium phosphate to retain cadmium. The mixing of hydrogen
            with  the  inert purge  gas  has also  been  used to  suppress
            chemical interference. The hydrogen acts as a reducing agent
            and aids in molecular dissociation.

                            7000A  -  3                       Revision  1
                                                            July 1992

-------
      3.    Analyze  the  sample  by  method  of  standard  additions  while
            noticing the precautions and limitations of its use (see Step
            8.7.2).

      3.2.2   Gases generated in the furnace during atomization may have
molecular absorption bands  encompassing the analytical  wavelength. When
this  occurs,  use  either  background correction  or choose  an  alternate
wavelength.  Background  correction  may  also compensate  for nonspecific
broad-band absorotion interferenre.
waveiengin.  oacKgrouna  correction
broad-band absorption interference
      3.2.3   Continuum background correction cannot correct for all types
of background  interference.  When the background  interference  cannot be
compensated for, chemically remove the analyte or use an alternate form of
background correction, e.g., Zeeman background correction.

      3.2.4     Interference from  a smoke-producing  sample matrix  can
sometimes  be  reduced  by  extending the   charring  time  at  a  higher
temperature or  utilizing  an ashing  cycle in the  presence of air.   Care
must be taken, however, to prevent loss of the analyte.

      3.2.5   Samples  containing large amounts of organic materials should
be oxidized  by conventional acid digestion before being placed  in  the
furnace. In this way, broad-band absorption will be minimized.

      3.2.6   Anion interference  studies  in  the graphite furnace indicate
that,  under  conditions  other  than isothermal,  the  nitrate anion  is
preferred.  Therefore, nitric  acid  is  preferable  for any  digestion or
solubilization  step.   If  another acid  in  addition to  nitric acid  is
required,  a minimum  amount should be used.  This applies particularly to
hydrochloric and, to  a lesser extent, to sulfuric and phosphoric acids.

      3.2.7   Carbide  formation resulting from the chemical environment of
the furnace has been observed. Molybdenum may be cited as an example. When
carbides form, the metal  is released very slowly  from the resulting metal
carbide  as atomization continues.  Molybdenum may  require 30 seconds or
more  atomization time  before the  signal   returns  to  baseline  levels.
Carbide  formation  is  greatly reduced and the sensitivity increased with
the  use of  pyrolytically coated graphite.  Elements that  readily form
carbides are noted with the  symbol  (p)  in Table  1.

      3.2.8    For comments  on spectral  interference, see Step 3.1.5.

      3.2.9   Cross-contamination and contamination of the sample can be
major sources of error because of the extreme  sensitivities achieved with
the furnace. The sample preparation  work  area  should  be kept  scrupulously
clean.  All  glassware  should be cleaned as  directed  in  Step 4.8.   Pipet
tips are a frequent source of contamination. If  suspected, they should be
acid  soaked with  1:5 nitric  acid  and  rinsed  thoroughly with  tap  and
reagent  water. The use of  a  better  grade of pipet tip can greatly reduce
this problem. Special attention should  be given  to reagent blanks  in both
analysis and  in  the  correction of analytical  results. Lastly, pyrolytic
graphite,  because of the production process  and  handling,  can  become
contaminated.  As  many as  five  to  ten  high-temperature  burns  may be
required to clean the tube before use.

                             7000A - 4                        Revision 1
                                                             July  1992

-------
4.0   APPARATUS AND MATERIALS

      4.1   Atomic  absorption spectrophotometer  -  Single-  or  dual-channel,
single- or double-beam instrument having a grating monochromator, photomultiplier
detector, adjustable slits, a wavelength  range  of 190 to 800 nm, and provisions
for interfacing with a graphical display.

      4.2   Burner  -  The  burner  recommended  by  the  particular  instrument
manufacturer should be used.  For certain  elements  the  nitrous  oxide burner is
required.

      4.3   Hollow  cathode  lamps  -  Single-element  lamps  are preferred  but
multielement lamps may be used. Electrodeless discharge lamps may also be used
when available.  Other types of lamps meeting the performance criteria of this
method may be used.

      4.4   Graphite  furnace  -  Any  furnace  device  capable  of  reaching  the
specified temperatures is satisfactory.

      4.5   Graphical  display and  recorder  -  A recorder is  recommended  for
furnace work so that there  will be a permanent record  and that any problems with
the analysis  such as drift,  incomplete  atomization, losses  during charring,
changes in sensitivity,  peak shape, etc., can be easily recognized.

      4.6   Pipets - Microliter,  with  disposable tips. Sizes can range from 5 to
100 uL  as  required.  Pipet  tips  should   be  checked  as  a possible  source  of
contamination  prior to  their  use.  The  accuracy of  automatic pipets  must  be
verified daily.  Class A pipets can  be  used for the measurement of volumes larger
than 1 ml_.

      4.7   Pressure-reducing valves  - The supplies of fuel and oxidant should
be maintained at pressures somewhat  higher  than the controlled operating pressure
of the instrument by suitable valves.

      4.8   Glassware -  All   glassware,  polypropylene,  or Teflon containers,
including sample bottles, flasks and pipets,  should be washed in the following
sequence:  detergent, tap  water, 1:1  nitric  acid,  tap  water,  1:1  hydrochloric
acid,  tap water, and reagent water. (Chromic  acid  should not  be  used  as  a
cleaning  agent for  glassware if chromium is to be included  in the analytical
scheme.)  If it can be documented through an active analytical  quality control
program  using  spiked samples  and   reagent  blanks  that  certain  steps  in  the
cleaning  procedure  are  not required  for  routine  samples, those  steps  may  be
eliminated from the procedure.


5.0   REAGENTS

      5.1  Reagent grade chemicals  shall  be used in all tests. Unless otherwise
indicated, it is intended that all  reagents shall conform  to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such specifications are available. Other grades  may be used, provided it is first
ascertained that the reagent  is  of  sufficiently  high purity  to permit its use


                                   7000A  - 5                       Revision 1
                                                                  July 1992

-------
without lessening the  accuracy  of the determination.  All  reagents  should be
analyzed to provide proof that all constituents are below the MDLs.

      5.2   Reagent  water.   All  references to water  in  this method  refer to
reagent water unless otherwise  specified.   Reagent grade water will  be  of at
least 16 Mega Ohm quality.

      5.3  Nitric acid  (concentrated),  HN03.  Use a spectrograde acid certified
for AA use.  Prepare  a 1:1 dilution with water by adding the concentrated acid to
an equal volume of water.  If the reagent blank is less than the IDL, the acid
may be used.

      5.4  Hydrochloric acid  (1:1),  HC1.  Use a spectrograde acid certified for
AA use. Prepare a 1:1 dilution with  water by adding the concentrated acid to an
equal volume of water.   If the reagent  blank is less than the IDL, the acid may
be used.

      5.5  Fuel  and  oxidant - High purity acetylene is generally acceptable. Air
may  be  supplied from  a compressed  air  line,  a  laboratory compressor,  or  a
cylinder of compressed air and should be clean and dry.   Nitrous oxide is also
required for certain determinations. Standard,  commercially  available argon and
nitrogen are required for furnace work.

      5.6 Stock standard metal solutions - Stock standard  solutions  are prepared
from  high   purity  metals, oxides,  or  nonhygroscopic  salts  using water  and
redistilled nitric or hydrochloric  acids.  (See individual methods for specific
instructions.)  Sulfuric or phosphoric acids should be avoided as they produce
an  adverse  effect  on  many   elements.  The stock solutions  are  prepared  at
concentrations  of  1,000  mg  of  the metal   per liter. Commercially  available
standard solutions may also be used.  Where the sample viscosity, surface tension,
and  components  cannot  be accurately  matched  with  standards,  the  method of
standard addition (MSA) may be used (see Step 8.7).

      5.7  Calibration standards  -  For those instruments which do not read out
directly  in  concentration,   a  calibration curve is  prepared  to cover  the
appropriate  concentration range.    Usually,   this  means  the  preparation  of
standards which produce an absorbance of 0.0 to 0.7.  Calibration standards are
prepared by diluting the stock metal solutions at the  time of analysis. For best
results, calibration standards  should  be  prepared fresh each time a batch of
samples is analyzed. Prepare  a blank and at  least  three calibration  standards in
graduated amounts in the appropriate range  of the  linear  part of the curve. The
calibration  standards   should be  prepared using  the same type  of acid or
combination of acids and at the same concentration as  will  result  in the samples
following processing. Beginning  with the blank and working toward the highest
standard, aspirate the solutions  and record the readings. Repeat the operation
with both the calibration standards  and the samples a  sufficient number of times
to secure a  reliable average reading for each solution. Calibration standards for
furnace procedures should be  prepared as described on the individual sheets for
that metal.  Calibration curves  are always required.
                                   7000A -  6                      Revision 1
                                                                  July 1992

-------
6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1  See the introductory material in Chapter Three,  Metallic Analytes.


7.0   PROCEDURE

      7.1  Preliminary treatment of waste water, ground water, EP extracts, and
industrial waste is always necessary because of the complexity and variability
of sample matrices. Solids, slurries, and suspended material  must be subjected
to a solubilization process before analysis.  This process may vary because of the
metals  to  be  determined  and  the  nature of  the  sample  being  analyzed.
Solubilization and digestion procedures are presented in Chapter Three, Step 3.2,
Sample Preparation Methods.   Samples which  are  to be analyzed  for dissolved
constituents need not be digested if they have been filtered and acidified.

      7.2  Direct aspiration (flame)  procedure

            7.2.1  Differences  between  the   various  makes  and  models  of
      satisfactory atomic absorption  spectrophotometers prevent the formulation
      of  detailed instructions  applicable  to every  instrument.  The  analyst
      should follow the manufacturer's operating instructions for a particular
      instrument.  In general, after choosing the proper lamp for the analysis,
      allow the lamp  to warm up for a minimum of 15 minutes, unless operated in
      a double-beam mode.  During this period, align  the instrument, position the
      monochromator at the correct wavelength,  select the proper monochromator
      slit  width,  and  adjust  the current  according to  the  manufacturer's
      recommendation. Subsequently, light  the  flame and regulate  the  flow of
      fuel and oxidant. Adjust  the burner  and  nebulizer  flow rate for maximum
      percent absorption and stability.  Balance the photometer. Run a series of
      standards of the element  under  analysis.  Construct a calibration curve by
      plotting the concentrations of the standards against absorbances. Set the
      curve corrector  of  a direct reading  instrument  to read out  the  proper
      concentration.    Aspirate the  samples and  determine  the  concentrations
      either directly or from the calibration curve. Standards must be run each
      time a sample or series of samples is run.

      7.3   Furnace procedure

            7.3.1  Furnace  devices (flameless  atomization) are  a most  useful
      means  of  extending  detection  limits.  Because  of differences  between
      various  makes   and  models  of  satisfactory  instruments,  no  detailed
      operating  instructions  can be given  for each instrument.  Instead,  the
      analyst should follow the instructions provided by  the manufacturer of a
      particular instrument.

            7.3.2   Background  correction  is  important  when using  flameless
      atomization, especially below 350  nm.  Certain samples, when atomized, may
      absorb or scatter light from the lamp. This can be caused by the presence
      of gaseous molecular species, salt particles,  or smoke  in the sample beam.
      If no correction  is made, sample absorbance will  be greater than it should
      be, and the analytical  result will be  erroneously high. Zeeman background
      correction is effective in overcoming composition or structured background


                                   7000A - 7                      Revision 1
                                                                  July 1992

-------
interferences.  It  is  particularly useful when  analyzing for As  in  the
presence of Al and when analyzing for Se in  the presence of Fe.

      7.3.3   Memory  effects  occur when  the  analyte  is  not  totally
volatilized during atomization. This condition depends on  several factors:
volatility  of  the  element  and   its  chemical  form,  whether  pyrolytic
graphite  is  used,  the  rate  of  atomization,  and  furnace design.  This
situation is detected through blank burns.  The tube should be cleaned by
operating the  furnace at  full  power for the required  time  period,  as
needed, at regular intervals during the series of determinations.

      7.3.4  Inject  a measured  micro!iter  aliquot of  sample  into  the
furnace  and  atomize.  If  the  concentration  found  is  greater  than  the
highest standard, the sample  should be diluted in the same acid matrix and
reanalyzed. The use of multiple injections can improve accuracy and help
detect furnace pipetting errors.

      7.3.5   To verify the  absence  of interference,  follow  the serial
dilution procedure given in Step 8.6.

      7.3.6  A check  standard should be run after approximately every 10
sample  injections.  Standards are  run  in part  to  monitor the  life  and
performance of the graphite tube. Lack of reproducibility or significant
change in the signal  for the standard  indicates that  the tube should be
replaced. Tube life depends on sample matrix and atomization temperature.
A  conservative  estimate would be  that  a  tube will   last  at  least  50
firings. A pyrolytic  coating will extend that estimated life by a factor
of three.

7.4   Calculation

      7.4.1  For determination of metal  concentration by direct aspiration
and furnace:  Read the metal  value from the calibration curve or directly
from the read-out system of the instrument.

      7.4.2   If dilution of sample was required:

    ug/L metal  in sample = A  (C_+_B)
                                C
where:

      A =   ug/L of metal  in diluted aliquot  from calibration curve.
      B =   Acid blank matrix used for dilution, mL.
      C =   Sample aliquot, mL.

      7.4.3   For solid  samples,  report all  concentrations in consistent
units based on wet weight. Hence:

   ug metal/kg  sample = A  x V
                           W
where:

      A =   ug/L of metal  in processed sample from  calibration  curve.
      V =   Final volume of  the processed sample, mL.
      W =   Weight of sample, grams.

                             7000A -  8                       Revision 1
                                                            July  1992

-------
            7.4.4  Different injection volumes  must not be used for samples and
      standards.  Instead,  the  sample should  be diluted  and  the same  size
      injection volume be used for both  samples  and  standards.  If dilution of
      the sample was required:

          ug/L of metal in sample = Z ( C + B)
                                         C
      where:

            Z =   ug/L of metal  read  from calibration curve or read-out system.
            B =   Acid blank matrix used for dilution mL.
            C =   Sample aliquot, ml.


8.0   QUALITY CONTROL

      8.1  All quality control data should be maintained and available for easy
reference or inspection.

      8.2  A  calibration  curve  must  be prepared each day with  a  minimum of a
calibration blank and three  standards.  After calibration, the calibration curve
must be verified by use of at  least a  calibration blank and a calibration check
standard (made from a reference material or other  independent standard material)
at or near the  mid-range. The calibration reference  standard  must be measured
within 10 % of  it's true value for the curve to be considered valid.

      8.3  If more  than  10  samples per day are analyzed,  the  working standard
curve must  be verified by measuring  satisfactorily  a mid-range  standard   or
reference standard after every 10 samples.  This sample value must be within 20%
of the true value, or the previous ten samples need to be reanalyzed.

      8.4  At least one matrix spike and one matrix spike duplicate sample shall
be included in each analytical batch.  A laboratory control sample  shall also be
processed with  each sample batch.  Refer to Chapter One for more information.

      8.5  Where the sample matrix is so complex that viscosity, surface tension,
and  components  cannot  be accurately  matched  with  standards,  the method  of
standard addition  (MSA)  is  recommended (see Section 8.7 below).   Section 8.6
provides tests  to evaluate the need for using the MSA.

8.6   Interference tests

      8.6.1  Dilution test - For each analytical batch select one typical sample
for  serial  dilution  to  determine  whether  interferences  are  present.    The
concentration of the analyte should be at least  25 times the estimated detection
limit.  Determine the apparent concentration in  the undiluted sample.  Dilute the
sample by a minimum of five  fold  (1+4)  and  reanalyze.  If all of the samples in
the batch are below  10  times  the detection limits,  perform  the  spike recovery
analysis described below.  Agreement within 10% between the concentration for the
undiluted  sample   and  five  times  the  concentration for  the diluted  sample
indicates the absence of  interferences, and such samples may be analyzed without
using the method of standard additions.
                                   7000A  -  9                       Revision 1
                                                                  July 1992

-------
      8.6.2  Recovery test - If results from the dilution test do not agree, a
matrix interference may be suspected and a spiked sample should be analyzed to
help confirm the finding from the dilution test.  Withdraw another aliquot of the
test sample and add a known amount of  analyte to bring the concentration of the
analyte to 2 to 5 times the original concentration.  If all of the samples in the
batch have analyte concentrations below the detection limit, spike the selected
sample at 20 times  the detection limit.  Analyze the spiked sample and calculate
the spike recovery.  If the recovery is less than 85% or greater than 115%, the
method of standard additions shall be used for all  samples in the batch.

      8.7   Method  of  standard  additions  - The  standard  addition  technique
involves  adding known  amounts of  standard  to one  or  more  aliquots  of the
processed sample solution. This technique compensates for a sample constituent
that enhances or depresses the analyte signal,  thus producing a different slope
from  that  of  the calibration  standards.  It  will  not  correct  for  additive
interferences which  cause  a  baseline  shift.  The method of standard additions
shall be used for analysis  of all  EP extracts,  on all analyses submitted as part
of a delisting petition, and whenever a new sample matrix is being analyzed.

            8.7.1  The simplest version of this technique is the single-addition
      method,  in which two identical  aliquots  of the  sample solution,  each of
      volume Vx, are taken. To the first (labeled A) is  added a  known volume Vs
      of a  standard analyte solution of concentration  Cs. To the second aliquot
      (labeled  B)  is added the same  volume Vs  of the  solvent.  The analytical
      signals of A and B are measured  and corrected for nonanalyte signals. The
      unknown  sample concentration Cx is calculated:
            C
             X
                      (SA-SB)VX

      where  SA  and  SB  are  the  analytical  signals  (corrected for the blank) of
      solutions A and  B, respectively. Vs and Cs should be chosen so that SA is
      roughly twice SB on the average, avoiding excess dilution of the sample.
      If a separation or concentration  step is used, the additions are best made
      first  and carried  through the entire procedure.

             8.7.2   Improved  results  can  be obtained by  employing  a series of
      standard  additions. To equal volumes of the  sample are added a series of
      standard  solutions containing different known quantities of the analyte,
      and  all  solutions are  diluted to  the  same final volume.  For example,
      addition  1  should be  prepared  so  that the resulting  concentration is
      approximately  50 percent  of the expected absorbance from the endogenous
      analyte in the  sample.  Additions 2 and 3 should  be prepared so that the
      concentrations  are approximately  100  and  150  percent of  the expected
      endogenous  sample  absorbance.  The  absorbance  of  each  solution  is
      determined and  then  plotted on  the vertical axis of  a  graph,  with the
      concentrations of the known  standards plotted on the horizontal  axis. When
      the  resulting  line  is extrapolated  to zero  absorbance,  the  point of
      interception  of  the  abscissa  is  the  endogenous  concentration  of the
      analyte in the sample. The abscissa on the  left of  the ordinate is scaled
      the  same  as  on  the  right  side,  but in  the  opposite  direction from the
                                   7000A  -  10                      Revision  1
                                                                  July  1992

-------
      ordinate.  An example  of a plot so obtained is  shown  in Figure 1.  A linear
      regression program may be used to obtain the  intercept  concentration.

            8.7.3  For  the  results of  this  MSA  technique  to  be  valid,  the
      following limitations must be taken into consideration:

            1.    The apparent  concentrations from the calibration curve must be
                  linear over the concentration range of concern.  For the best
                  results,  the slope of the MSA plot should  be nearly the same
                  as  the  slope  of  the   standard  curve.  If  the  slope  is
                  significantly different (greater  than 20%), caution should be
                  exercised.

            2.    The effect of the  interference should not vary  as the ratio of
                  analyte  concentration  to  sample  matrix  changes,  and  the
                  standard addition should respond  in a  similar manner as the
                  analyte.

            3.    The determination must be free of spectral  interference and
                  corrected for nonspecific background interference.

      8.8   All  quality  control  measures described  in  Chapter  One  should  be
followed.
9.0  METHOD PERFORMANCE

      9.1  See individual methods.


10.0 REFERENCES

1.    Methods for  Chemical  Analysis of  Water  and Wastes;  U.S.  Environmental
Protection Agency.  Office of Research and Development. Environmental Monitoring
and Support  Laboratory.  ORD Publication Offices  of Center  for  Environmental
Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.

2.    Rohrbough, W.G.;  et  al.  Reagent  Chemicals,  American  Chemical  Society
Specifications,  7th ed.; American Chemical  Society: Washington, DC, 1986.

3.    1985 Annual Book of ASTM Standards, Vol. 11.01; "Standard Specification for
Reagent Water";  ASTM: Philadelphia, PA, 1985; D1193-77.
                                  7000A - 11                      Revision 1
                                                                  July 1992

-------
                                  TABLE 1.
                   ATOMIC ABSORPTION CONCENTRATION RANGES
Metal
Aluminum
Antimony
Arsenic
Barium
Beryllium
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Lithium
Magnesium
Manganese
Mercury
Molybdenum(p)
Nickel
Osmium
Potassium
Selenium
Silver
Sodium
Strontium
Thallium
Tin
Vanadium(p)
Zinc

Detection Limit
(mg/L)
0.1
0.2
0.002
0.1
0.005
0.005
0.01
0.05
0.05
0.02
0.03
0.1
0.002
0.001
0.01
0.0002
0.1
0.04
0.03
0.01
0.002
0.01
0.002
0.03
0.1
0.8
0.2
0.005

Sensitivity
(mg/L)
1
0.5
0.4
0.025
0.025
0.08
0.25
0.2
0.1
0.12
0.5
0.04
0.007
0.05
0.4
0.15
1
0.04
0.06
0.015
0.15
0.5
4
0.8
0.02
Furnace Procedure4'0
Detection Limit
(ug/L)
..
3
1
2
0.2
0.1
--
1
1
1
1
1
--
0.2
1
--
2
0.2
--
--
1
--
4
0.05
                                                                                   i
      NOTE:
The symbol (p) indicates the use of  pyrolytic  graphite with
the furnace procedure.
aFor furnace sensitivity values,  consult instrument  operating  manual.

Gaseous hydride method.

'ihe listed furnace values are those expected when using a 20-uL injection and
normal  gas  flow, except in the cases of arsenic and selenium, where gas interrupt
is used.

      vapor technique.
                                  7000A - 12
                                                Revision 1
                                                July 1992

-------
                                FIGURE  1.
                         STANDARD ADDITION PLOT
               CD
               o
               CO
               .a
               o
               CO
  Zero
Absorbance
                                                            Concentration
      Cone, of     Addn 0      Addn 1       Addn 2      Addn 3
      Sample      No Addn    Addn of 50%  Addn of 100% Addn of 150%
                             of Expected   of Expected   of Expected
                             Amount      Amount      Amount
                               7000A - 13
Revision 1
July 1992

-------
                                     METHOD 7000A
                            ATOMIC ABSORPTION  METHODS
                         c
                         7.1 Solubili2«
                           and digeftt
                           •ample  (tee
                           Chapter 3,
                          Section  3  2)
 7.2 1 Choo««
  and prepare
  hollow tub*
 cathode lamp
                          7.3.1 Follow
                            operating
                          in»tructlon*
                         fron instrument
                          manufacturer
 7.2.1 ftdjust
   and align
   equipment
  7.2.1  Light
   flan* and
   regulate
   7.21  Run
   •tandard*
  7.3.3  Clean
     tub*
                           732 Hake
                           background
                           correction
7 3.4  Inject
 and atomize
   part of
   (ample
7 2.1 Construct
  calibration
 curve and  *et
curve corrector
     7.2  1
   A*pirate
    •ample
                                                7.3 4 Dilutfl
                                                  • atnpl*
                            7  3.5 U.«
                          intarf«r«nc«
                         t«»t» to varify
                           abvanca of
                          interfarBnc*
  7.2.1  Run a
     check
   itandard
 7 4 Determine
concentration!
 7 3  6 Run
   check
  •tandard
                              Stop
                                           7000A  -  14
                                                             Revision  1
                                                             July 1992

-------
                                 METHOD 7020

               ALUMINUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Aluminum may be as much  as 15% ionized in a nitrous-oxide/acetylene
flame.  Use of an ionization  suppressor  (1,000  ug/mL K as KC1)  as in Method
7000, Paragraph 3.1.4, will eliminate this interference.

     3.3  Aluminum is a  very  common  contaminant,  and  great care should be
taken to avoid contamination.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Aluminum hollow cathode lamp.
          4.2.2  Wavelength:  324.7 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Nitrous oxide.
          4.2.5  Type of flame:  Fuel rich.
          4.2.6  Background correction:  Not required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards:

          5.2.1  Stock solution:  Dissolve  1.000 g of aluminum metal in dilute
     HC1 with gentle warming.  Dilute to 1  liter with Type  II water.  Alterna-
     tively, procure  a  certified  standard  from  a   supplier  and verify by
     comparison with a second standard.
                                  7020 - 1
                                                         Revision
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions   of  the   stock   solution  to  be   used as
     calibration standards at the time of analysis.   The  calibration standards
     should  be  prepared  using  the  same   type  of  acid  and   at   the  same
     concentration  as  will   result  in  the  sample  to be  analyzed after
     processing.  Samples and standards  should also  contain 2 ml KC1/100 ml
     solution (Paragraph 3.2  above).

     5.3  Potassium chloride  solution;  Dissolve 95  g potassium chloride (KC1)
in Type II water and dilute to 1 liter.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3,  Sample Handling  and  Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation  of the sample
are given in Chapter Three, Section 3.2.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  Precision and accuracy data are available in Method 202.1 of Methods
for Chemical Analysis of Water and Wastes.

     9.2  The   performance  characteristics  for  an  aqueous  sample  free of
interferences are:

     Optimum concentration range:   5-50 mg/L, with a wavelength of 309.3 nm.
     Sensitivity:   1 mg/L.
     Detection  limit:   0.1 mg/L.


10.0   REFERENCES

1.   Methods for  Chemical   Analysis   of  Water   and Wastes, EPA-600/4-82-055,
Method 202.1,  December 1982.
                                   7020 - 2
                                                          Revision
                                                          Date   September 1986

-------
                 METHOD 7020

ALUMINUM (ATOMIC  ABSORPTION. DIRECT ASPIRATION)
                 5.O
                     Prepare
                    standards
                 7.1
                      For
                     sample
                preparation see
                   chapter 3.
                   section 3.2
                 7.2
                 Analyze using
                 Method 700O.
                 Section 7.2
              f     Stop      J
             7020 - 3
                                       Revision       o
                                       Date  September  1986

-------
                                 METHOD 7040

               ANTIMONY (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  In the presence of  lead  (1,000  mg/L), a spectral interference may
occur at the 217.6-nm resonance  line.    In  this case, the 231.1-nm antimony
line should be used.

     3.3  Increasing   the   acid   concentrations   decreases   the  antimony
absorption.  To avoid this effect,  the  acid concentration in the samples and
in the standards should be matched.

     3.4  Excess concentrations  of  copper  and  nickel  (and  possibly other
elements), as well as acids,  can  interfere  with  antimony analyses.  If the
sample contains these matrix types, either matrices of the standards should be
matched to those of  the  sample  or  the  sample  should  be analyzed using a
nitrous oxide/acetylene flame.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Antimony hollow cathode lamp or electrode!ess discharge lamp.
          4.2.2  Wavelength:  217.6 nm  (primary); 231.1 nm (secondary).
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Fuel lean.
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See  Section 5.0 of Method 7000.
                                   7040 -  1
                                                         Revision      0
                                                         Date  September 1986

-------
     5.2   Preparation  of standards;

          5.2.1   Stock  solution:     Carefully  weigh  2.7426  g  of  antimony
     potassium tartrate,  K(SbO)C4H406* 1/2^0   (analytical  reagent grade), and
     dissolve in Type  II water.  Dilute  to  1  liter with Type  II water; 1 ml =
     1  mg Sb (1,000 mg/L).   Alternatively, procure a certified standard from a
     supplier and verify by comparison with a  second standard.

          5.2.2   Prepare  dilutions   of   the   stock  solution  to  be  used as
     calibration standards at the  time of analysis.  The calibration standards
     should contain 0.2% (v/v) HN03  and   1-2%  v/v HC1, prepared  using the same
     types of acid and  at  the  same concentrations  as  in  the  sample after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3,  Sample  Handling  and  Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Method 3005.    Method  3005,  a soft  digestion,  is presently the
only digestion procedure recommended for Sb.  It yields better recoveries than
either Method 3010 or Method 3050. There  1s  no hard digestion  for Sb at this
time.

     7.2  See Method 7000,  Paragraph 7.2, Direct Aspiration Procedure.


8.0  QUALITY CONTROL

     8.1  See Section 8.0  of Method 7000.
9.0  METHOD  PERFORMANCE

     9.1  The  performance  characteristics  for  an  aqueous  sample  free of
Interferences are:

     Optimum concentration  range:   1-40 mg/L with a wavelength of 217.6 nm.
     Sensitivity:  0.5 mg/L.
     Detection limit:  0.2  mg/L.

     9.2  In a   single   laboratory,  analysis  of  a mixed industrial-domestic
waste  effluent,  digested with Method 3010,  at concentrations of 5.0 and 15 mg
Sb/L gave the standard deviations of +0.08 and +0.1, respectively.  Recoveries
at these levels  were  96% and 97%, respectively.

     9.3  For  concentrations   of   antimony   below  0.35  mg/L,  the  furnace
procedure  (Method  7041)  is  recommended.


                                  7040 -  2
                                                         Revision      0
                                                         Date  September 1986

-------
10.0  REFERENCES

1.  Methods  for  Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December 1982, Method 204.1.
                                  7040 - 3
                                                         Revision
                                                         Date  September 1986

-------
                 METHOD 70*0
ANTIMONY  (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
                I      Start      J
5
.0

Prepare
standards
7.1
prepar
ch
sect!
For
sample
ation-
apter
on 3.1
sea
3.
.3
                 7.Z
                 Analyza using
                  Method 7000.
                  Section 7.a
               f     Stop      J
               7040 - 4
                                         Revision       0
                                         Date  September  1986

-------
                                 METHOD 7041

               ANTIMONY (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 1f interferences are suspected.

     3.2  High lead concentration may cause a measurable spectral  interference
on the  217.6-nm  line.    If  this  interference  is  expected, the secondary
wavelength should be employed or Zeeman background correction used.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Drying time and temp:  30 sec at 125*C.
          4.2.2  Ashing time and temp:  30 sec at 800'C.
          4.2.3  Atomizing time and temp:  10 sec at 2700*C.
          4.2.4  Purge gas:  Argon or nitrogen.
          4.2.5  Wavelength:  217.6 nm  (primary); 231.1 nm  (alternate).
          4.2.6  Background correction:  Required.
          4.2.7  Other operating parameters should be  set as specified by the
     .particular instrument manufacturer.
     NOTE:   The above concentration values and instrument conditions are for a
             Perkin-Elmer  HGA-2100, based  on  the  use  of  a 20-uL Injection,
             continuous-flow purge  gas,  and  nonpyrolytic  graphite.  Smaller
             sizes of  furnace  devices  or  those  employing  faster  rates of
             atomization can be  operated  using lower atomization temperatures
             for shorter time periods than the above-recommended settings.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.
                                  7041 - 1
                                                         Revision
                                                         Date  September 1986

-------
     5.2  Preparation of standards;

          5.2.1  Stock  solution:     Carefully  weigh   2.7426   g   of   antimony
     potassium tartrate (analytical   reagent  grade)   and   dissolve in Type  II
     water.   Dilute  to  1  liter  with  Type  II  water;   1   mL  =   1  mg  Sb
     (1,000  mg/L).    Alternatively,  procure  a  certified  standard  from a
     supplier and verify by comparison with a second  standard.

          5.2.2  Prepare  dilutions   of  the  stock  solution   to  be  used  as
     calibration standards at the time of analysis.  The calibration  standards
     should contain 0.2% (v/v)  HN03  and  1-2%  (v/v) HCl, prepared  using the
     same types of acid and at the  same concentrations as  in  the  sample after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample  Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparat1on;  The  procedures  for  preparation of  the sample
are given in Method 3005.Method  3005,  a soft  digestion,  is presently the
only digestion procedure recommended for Sb.  It yields better recoveries than
either Method 3010 or Method 3050.  There  1s no hard digestion for  Sb at this
time.
     NOTE:  The addition of HCl  acid  to  the  digestate prevents the furnace
            analysis of this digestate for many other metals.

     7.2  See Method 7000, Paragraph 7.3,  Furnace Procedure.   The calculation
is given in Method 7000, Paragraph 7.4.


8.0  QUALITY CONTROL

     8.1  See Section  8.0  of Method  7000.


9.0  METHOD  PERFORMANCE

     9.1   Precision  and  accuracy data  are  not  available at this time.

     9.2   The   performance  characteristics   for  an   aqueous  sample  free  of
interferences  are:

     Optimum concentration range:   20-300  ug/L.
     Detection  limit:   3 ug/L.
                                   7041 - 2
                                                          Revision
                                                          Date  September 1986

-------
10.0  REFERENCES

1.   Methods for  Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December 1982, Method 204.2.
                                   7041 -  3
                                                         Revision
                                                         Date  September  1986

-------
                 METHOD 7O41

ANTIMONY (ATOMIC ABSORPTION.  FURNACE  TECHNIQUE)
              f     Start     j
                5.0
                    Prepare
                   standards
7.1
prepar
ct
sec
For
cample
•atton seo
apter -3.
tlon 3.2
                7.2
                Analyze  using
                 Method  7000.
                 Section 7.3.
               calculation 7.4
              f      Stop       J
             7041 -  4
                                       Revision       0
                                       Date  September 1986

-------
                                 METHOD 7060A

                ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)
1.0  SCOPE AND APPLICATION

      1.1   Method  7060  is  an  atomic  absorption  procedure  approved  for
determining the concentration of arsenic in wastes,  mobility procedure extracts,
soils,  and  ground water.   All  samples  must  be  subjected to  an  appropriate
dissolution step prior to analysis.

2.0  SUMMARY OF METHOD

      2.1   Prior to analysis by  Method 7060,  samples must  be prepared in order
to convert  organic forms of  arsenic  to  inorganic forms,  to  minimize organic
interferences, and to convert the  sample  to a suitable solution for analysis.
The sample preparation procedure varies depending on the sample matrix.  Aqueous
samples are subjected to the acid digestion  procedure described  in this method.
Sludge samples are prepared using the procedure described  in Method 3050.

      2.2  Following the appropriate dissolution of the sample, a representative
aliquot of the digestate is  spiked  with a  nickel nitrate solution and is placed
manually or by means of  an automatic sampler into a graphite tube furnace.  The
sample  aliquot  is  then  slowly  evaporated  to dryness,  charred  (ashed),  and
atomized.  The absorption of hollow cathode  or EDL radiation during atomization
will  be proportional to  the  arsenic concentration.  Other modifiers may be used
in  place  of  nickel  nitrate if   the analyst  documents the  chemical  and
concentration used.

      2.3   The typical  detection limit for water samples  using this method is
1 ug/L.  This detection limit may not be achievable when  analyzing waste samples.

3.0  INTERFERENCES

      3.1   Elemental arsenic and many of its compounds  are  volatile; therefore,
samples may be subject to losses of arsenic during sample  preparation.   Spike
samples  and  relevant  standard  reference  materials  should  be processed  to
determine if the chosen  dissolution method  is appropriate.

      3.2   Likewise,  caution  must  be  employed  during  the  selection  of
temperature and times for the dry and char (ash) cycles.  A  matrix modifier such
as nickel  nitrate   must be added to all  digestates prior to  analysis to minimize
volatilization losses during drying and ashing.

      3.3   In addition  to the normal interferences experienced during graphite
furnace analysis,  arsenic analysis can suffer from severe nonspecific absorption
and light scattering caused  by matrix  components during  atomization.   Arsenic
analysis  is  particularly susceptible  to these  problems  because  of its  low
analytical wavelength (193.7 nm).   Simultaneous background correction  must be
employed  to  avoid erroneously high results.   Aluminum  is a severe  positive
interferent  in  the analysis  of  arsenic,  especially  using D2  arc  background


                                   7060A - 1                      Revision 1
                                                                  September 1994

-------
correction.   Although  Zeeman background  correction  is  very useful  in this
situation, use of any appropriate background correction technique is acceptable.

      3.4   If the analyte  is not  completely  volatilized and removed from the
furnace during atomization,  memory effects will occur.    If  this  situation is
detected by means of blank  burns,  the  tube should  be  cleaned by operating the
furnace at full power at regular intervals in the analytical  scheme.

4.0  APPARATUS AND MATERIALS

      4.1   Griffin beaker or equivalent:  250 ml.

      4.2   Class A Volumetric flasks:   10-mL.

      4.3   Atomic absorption spectrophotometer:  Single or dual channel, single-
or  double-beam instrument  having   a  grating monochromator,  photo-multiplier
detector, adjustable slits,  a wavelength range of 190  to  800 nm, and provisions
for simultaneous  background correction and interfacing with a suitable recording
device.

      4.4   Arsenic hollow cathode  lamp,  or electrodeless discharge lamp  (EDL):
EDLs provide better sensitivity for arsenic analysis.

      4.5   Graphite furnace:  Any  graphite furnace device with the appropriate
temperature and timing controls.

      4.6   Data  systems  recorder:  A recorder  is  strongly  recommended for
furnace work so that there will be a permanent record and so that any problems
with the analysis  such as drift, incomplete atomization, losses during charring,
changes in sensitivity, etc., can easily be recognized.

      4.7   Pipets:  Microliter with disposable tips.   Sizes can range from
5 to 1,000 uL, as required.

5.0  REAGENTS

      5.1   Reagent water:  Water should be monitored for impurities.
All references to water will refer to reagent water.

      5.2   Concentrated nitric acid: Acid should be analyzed to determine  levels
of impurities.  If a method  blank using the acid  is 
-------
      5.5   Nickel nitrate solution (5%):  Dissolve 24.780 g of ACS reagent grade
Ni(N03)26H20 or equivalent in reagent water and dilute to 100 mL.

      5.6   Nickel nitrate solution (1%):   Dilute 20 ml of the 5% nickel nitrate
to 100 mL with reagent water.

      5.7   Arsenic working standards:  Prepare dilutions of  the stock solution
to be  used as calibration  standards at  the  time of the  analysis.   Withdraw
appropriate aliquots of the stock solution, add concentrated  HN03, 30% H202, and
5% nickel nitrate solution or other appropriate matrix modifier.   Amounts added
should be representative of the concentrations found in  the samples.  Dilute to
100 ml with reagent water.

6.0  SAMPLE COLLECTION, PRESERVATION, AND  HANDLING

      6.1   All  samples  must have  been  collected using a sampling  plan that
addresses the considerations discussed in  Chapter Nine of this manual.

      6.2   All sample containers must be  prewashed  with  detergents, acids, and
reagent water.  Plastic and glass containers are both suitable.

      6.3   Special  containers  (e.g.,  containers  used  for  volatile  organic
analysis) may  have to be  used if  very  volatile  arsenic compounds are  to be
analyzed.

      6.4   Aqueous samples must be acidified to a pH of <2 with nitric acid and
refrigerated prior to analysis.

      6.5   Although waste samples do not need to be refrigerated sample handling
and storage must comply with the minimum requirements established in Chapter One.

7.0  PROCEDURE

      7.1   Sample preparation:  Aqueous samples should be prepared in the manner
described in  Paragraphs  7.1.1-7.1.3.  Sludge-type  samples should be  prepared
according to Method 3050A.   The  applicability of a sample-preparation technique
to a new matrix  type must be demonstrated  by  analyzing  spiked samples  and/or
relevant standard reference materials.

            7.1.1    Transfer  a known volume of well-mixed sample to  a  250-mL
      Griffin  beaker or  equivalent;  add  2  mL  of  30% H202  and  sufficient
      concentrated HN03 to result in an acid concentration of 1%  (v/v).   Heat,
      until  digestion is  complete,  at 95°C  or until the volume is  slightly less
      than 50 mL.

            7.1.2   Cool, transfer to a volumetric flask, and bring back to 50
      mL with reagent water.

            7.1.3   Pipet 5 mL of this digested solution into  a 10-mL volumetric
      flask, add  1  mL  of the 1% nickel nitrate solution or  other appropriate
      matrix modifier, and dilute to  10 mL with reagent  water.   The  sample is
      now ready for injection into the furnace.


                                   7060A -  3                       Revision 1
                                                                  September  1994

-------
      7.2   The 193.7-nm wavelength line and a  background correction system are
required.  Follow the manufacturer's suggestions for all other spectrophotometer
parameters.

      7.3   Furnace parameters suggested by the manufacturer should be employed
as  guidelines.     Because   temperature-sensing   mechanisms   and  temperature
controllers can  vary between  instruments  or with  time,  the validity  of the
furnace parameters  must  be periodically confirmed by systematically altering the
furnace parameters  while analyzing a standard.  In this manner,  losses of analyte
due to overly high temperature settings or losses in sensitivity due to less than
optimum settings can be minimized.  Similar verification of furnace parameters
may be required for complex sample matrices.

      7.4   Inject a measured microliter aliquot  of sample  into the furnace and
atomize.  If the concentration found is greater than the highest standard, the
sample should  be diluted in  the same acid matrix  and reanalyzed.   The use of
multiple  injections  can  improve  accuracy  and  help detect  furnace  pipetting
errors.

8.0  QUALITY CONTROL

      8.1  Refer to section 8.0 of Method 7000.

9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data  are available in Method 206.2 of Methods
for Chemical Analysis of Water and Wastes.

      9.2  The optimal concentration range  for aqueous samples using this method
is 5-100  ug/L.   Concentration  ranges  for  non-aqueous samples will  vary with
matrix type.

      9.3   The data shown in  Table  1  were obtained from  records of state and
contractor laboratories.   The data are intended to  show the  precision of the
combined sample preparation and analysis method.

10.0  REFERENCES

1.    Methods  for  Chemical  Analysis of Water and  Wastes,  EPA-600/4-82-055,
December 1982,  Method 206.2.

2.    Gaskill,  A.,  Compilation and Evaluation of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075, September 1986.
                                   7060A -  4                       Revision 1
                                                                  September 1994

-------
                       TABLE  1. METHOD  PERFORMANCE  DATA
Sample                       Preparation                 Laboratory
Matrix                        Method                    Replicates
Contaminated soil               3050                   2.0,  1.8 ug/g

Oily soil                       3050                   3.3,  3.8 ug/g

NBS SRM 1646 Estuarine sediment 3050                   8.1,  8.33 ug/ga

Emission control dust           3050                   430,  350 ug/g


aBias of -30 and -28% from expected, respectively.
                                   7060A  -  5                       Revision 1
                                                                  September 1994

-------
                             METHOD  7060A
       ARSENIC (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)
7.1.1 Tranafer
   •aaple to
beaker,add H.O.
and cane. HNO.,
     heat
                                        7.1 Prepare
                                          aaaplea
                                       according to
                                       Method 3050
 7.1.2 Cool
and bring to
 voluM with
reagent water
  7.1.3  Pipet
 aolution into
  fla«k, add
nickel nitrate,
    dilute
                    7.2 Set up
                    inatruaent
                    operating
                    parameter
                       7.3
                   Periodically
                     verify
                     furnace
                   paraa>eter*
                    7.4 Inject
                    aliquot of
                    aaaple into
                    furnace,
                     atoaite
                   7.4 Beeord A»
                   eeneentration
                                                              7 4 Dilute
                                                              aajiple and
                                                               reanalyze
                      Stop
                               7060A  - 6
                                                                        Revision  1
                                                                        September 1994

-------
                                 METHOD 7061A

                 ARSENIC (ATOMIC ABSORPTION, GASEOUS HYDRIDE)
1.0  SCOPE AND APPLICATION

      1.1   Method 7061 is an  atomic  absorption  procedure for determining the
concentration of  arsenic  in wastes,  mobility  procedure  extracts,  soils,  and
ground water.   Method  7061A  is approved only for sample  matrices  that do not
contain high concentrations of chromium, copper, mercury, nickel, silver, cobalt,
and molybdenum.  All samples must be subjected to an appropriate dissolution step
prior to analysis.  Spiked samples and relevant  standard reference materials are
employed to determine the applicability of the method to a given waste.


2.0  SUMMARY OF METHOD

      2.1   Samples are prepared according to the nitric/sulfuric acid digestion
procedure  described  in this  method  (Step  7.1).   Next,   the arsenic  in  the
digestate  is reduced to the  trivalent form with tin  chloride.   The trivalent
arsenic is then converted to a volatile hydride using hydrogen produced from a
zinc/hydrochloric acid reaction.

      2.2   The volatile hydride is swept into an argon-hydrogen flame located
in the optical  path of  an  atomic absorption  spectrophotometer.   The resulting
absorption of the lamp radiation is proportional to the arsenic concentration.

      2.3   The typical detection limit for this method is 0.002 mg/L.


3.0  INTERFERENCES

      3.1   High concentrations of chromium, cobalt, copper, mercury, molybdenum,
nickel, and silver can cause analytical interferences.

      3.2   Traces of nitric acid left following the sample work-up can result
in analytical interferences.   Nitric acid must be distilled off by heating the
sample until fumes of sulfur trioxide (S03)  are observed.

      3.3   Elemental  arsenic and many of its compounds are volatile; therefore,
certain samples may be subject to losses of arsenic during sample preparation.


4.0  APPARATUS AND MATERIALS

      4.1   Beaker or equivalent - 100-mL.

      4.2   Electric  hot  plate  or  equivalent   -  adjustable  and  capable  of
maintaining a temperature of 90-95°C.
                                   7061A  -  1                       Revision 1
                                                                  July 1992

-------
            4.3.1   Medicine dropper  - Capable of fitting into a size "0" rubber
      stopper and delivering 1.5 ml.

            4.3.2   Pear-shaped reaction flask  -  50-mL, with two  14/20  necks
      (Scientific Glass JM-5835 or equivalent).

            4.3.3   Gas inlet-outlet  tube - Constructed from a micro cold-finger
      condenser (JM-3325) by cutting the portion  below the  14/20 ground-glass
      joint.

            4.3.4   Magnetic stirrer -  To homogenize the zinc slurry.

            4.3.5   Polyethylene  drying  tube -  10-cm,  filled  with glass  to
      prevent particulate matter from entering the burner.

            4.3.6   Flow meter - Capable of measuring 1 liter/min.

            4.3.7   Class A volumetric  flasks.

            4.3.8   Graduated cylinder  or equivalent.

      4.4   Atomic absorption spectrophotometer - Single or dual channel, single-
or  double-beam instrument  having a grating monochromator,  photo-multiplier
detector, adjustable slits,  a wavelength  range of  190 to 800 nm,  and provisions
for interfacing with a strip-chart recorder.

      4.5   Burner - Recommended by the particular instrument manufacturer for
the argon-hydrogen flame.

      4.6   Arsenic hollow cathode lamp or arsenic  electrodeless discharge lamp.

      4.7   Strip-chart recorder.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all reagents  shall  conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical  Society, where
such specifications are available. Other grades may be used,  provided it  is first
ascertained that the reagent  is  of sufficiently high purity to  permit  its use
without lessening the  accuracy of the determination.

      5.2   Reagent  Water.    Reagent water will  be  interferent  free.    All
references  to water  in  the method  refer  to  reagent water  unless otherwise
specified.

      5.3   Nitric  acid  (concentrated),  HNO,.    Acid  should  be  analyzed  to
determine levels  of  impurities.   If a  methoa blank  is < MDL, the  acid can be
used.
                                   7061A - 2                      Revision 1
                                                                  July 1992

-------
      5.4   Sulfuric acid  (concentrated),  H2S04.   Acid should  be  analyzed to
determine levels of  impurities.   If a method blank is < MDL,  the  acid can be
used.

      5.5   Hydrochloric acid (concentrated), HC1.   Acid should be analyzed to
determine levels of  impurities.   If a method blank is < MDL,  the  acid can be
used.

      5.6   Diluent - Add 100 ml 18N H2SO, and 400 ml concentrated HC1 to 400 ml
water and dilute to a final volume of 1 liter with water.

      5.7   Potassium iodide solution - Dissolve 20 g KI in 100 mL water.

      5.8   Stannous  chloride  solution  -  Dissolve  100  g  SnCl2  in  100  ml
concentrated HC1.

      5.9   Arsenic solutions

            5.9.1  Arsenic standard solution (1,000 mg/L)  -  Either procure a
      certified aqueous standard from a supplier and verify by comparison with
      a second standard, or dissolve 1.320 g of arsenic trioxide As203 in 100 ml
      of  water  containing  4  g  NaOH.    Acidify  the  solution   with 20  ml
      concentrated HN03 and dilute to 1  liter.

            5.9.2   Intermediate arsenic solution  -  Pipet  1 ml  stock arsenic
      solution  into  a  100-mL volumetric flask and bring to  volume with water
      containing 1.5 ml concentrated HN03/liter (1  ml  = 10  ug As).

            5.9.3  Standard arsenic  solution - Pipet  10 ml  intermediate arsenic
      solution  into  a  100-mL volumetric flask and bring to  volume with water
      containing 1.5 mL concentrated HN03/liter (1  mL  = 1 ug As).


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples must have  been  collected  using a  sampling plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with detergents, acids, and
water.  Plastic and glass  containers are both suitable.

      6.3   Special  containers  (e.g.  containers  used  for volatile  organic
analysis) may  have to  be  used if  very  volatile arsenic compounds are  to be
analyzed.

      6.4   Aqueous samples must be  acidified to a pH of < 2 with  nitric acid.

      6.5   Nonaqueous samples shall be refrigerated, when possible,  and  analyzed
as soon as possible.
                                   7061A -  3                       Revision 1
                                                                  July 1992

-------
7.0  PROCEDURE

      7.1   Place a 50-mL aliquot of digested sample  (or, in the case of analysis
of EP extracts, 50 ml) of the material to be analyzed in a 100-mL beaker.  Add
10 ml concentrated HNO, and 12 mL 18N H2S04.   Evaporate the sample in the hood
on an electric hot plate until white SO, fumes are observed (a volume of about
20 mL).  Do not let the sample char.  If charring occurs, immediately turn off
the heat, cool, and add an  additional  3  ml  of HN03.  Continue to add additional
HN03 in order  to  maintain  an excess  (as  evidenced  by the  formation  of brown
fumes).  Do not let the solution  darken,  because arsenic may be  reduced and lost.
When the sample remains colorless or straw yellow during evolution of S03 fumes,
the digestion is complete.  Cool the  sample,  add  about 25 ml  water,  and again
evaporate  until S03 fumes  are produced  in order to expel  oxides of nitrogen.
Cool.  Transfer the digested sample  to a 100-mL volumetric  flask.  Add 40 ml of
concentrated HC1 and bring to volume with water.

      7.2   Prepare  working  standards  from  the standard arsenic  solution.
Transfer 0,  0.5,  1.0, 1.5, 2.0, and  2.5 mL of standard  to 100-mL volumetric
flasks and bring to volume with diluent.  These concentrations  will be 0, 5, 10,
15, 20, and 25 ug As/liter.

      7.3   If EP  extracts  are  being  analyzed or if  a matrix interference is
encountered, take  the 15-,  20-, and  25-mg/liter  standards and quantitatively
transfer 25 mL  of each  of these standards into separate 50-mL volumetric flasks.
Add  10  mL  of the  prepared  sample to  each  flask.    Bring  to volume  with water
containing 1.5 mL HCl/liter.

      7.4   Add 10 mL of prepared sample to a 50-mL  volumetric  flask.  Bring to
volume  with  water  containing 1.5  mL HCl/liter.    This  is the  zero  addition
aliquot.

NOTE:       The absorbance from  the zero addition aliquot will  be  one-fifth that
            produced  by the prepared  sample.  The  absorbance  from the spiked
            samples will be one-half  that produced by  the standards  plus the
            contribution from one-fifth of the prepared sample.  Keeping these
            absorbances in mind will  assist in judging the  correct dilutions to
            produce optimum absorbance.

      7.5   Transfer a 25-mL  portion of the digested  sample or standard to the
reaction vessel and add 1  mL KI  solution.  Add 0.5 mL  SnCl2  solution.  Allow at
least  10  minutes  for the metal  to  be reduced to its lowest  oxidation state.
Attach the reaction vessel  to the special gas inlet-outlet  glassware.  Fill the
medicine dropper with 1.50 mL zinc  slurry that has been kept in suspension with
the magnetic stirrer.   Firmly  insert the stopper containing the  medicine dropper
into the side  neck of the  reaction  vessel.  Squeeze the bulb to introduce the
zinc slurry into the sample or standard solution.  The metal hydride will produce
a peak almost immediately.   After the  recorder pen begins  to return to the base
line, the  reaction vessel can be removed.

      CAUTION:    Arsine is very toxic.   Precautions must be  taken  to avoid
                   inhaling  arsine gas.
                                   7061A - 4                      Revision 1
                                                                  July  1992

-------
      7.6   Use  the  193.7-nm  wavelength and  background  correction  for  the
analysis of arsenic.

      7.7   Follow the  manufacturer's  instructions  for  operating an  argon-
hydrogen flame.   The  argon-hydrogen flame is colorless; therefore, it  may be
useful to aspirate a  low concentration of sodium to ensure  that  ignition  has
occurred.

      7.8   If  the  method  of  standard additions  was  employed,  plot  the
absorbances of spiked samples and blank vs. the concentrations. The extrapolated
value will  be one-fifth the concentration of the original sample.   If the plot
does not result in a straight line, a nonlinear interference is present.  This
problem can sometimes be overcome by dilution or addition of other reagents if
there is some  knowledge about the waste.  If the method of standard additions was
not required,  then the concentration can be part of the calibration curve.


8.0  QUALITY CONTROL

      8.1  Refer to section 8.0 of Method 7000.
9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are available in Method 206.3 of Methods
for Chemical Analysis of Water and Wastes.


10.0  REFERENCES

1.    Methods  For  Chemical  Analysis  of  Water  and Wastes,  EPA-600/4-82-055,
December 1982, Method 206.3.

2.    Rohrbough, W.G.;  et  al.  Reagent  Chemicals,  American Chemical  Society
Specifications, 7th ed.; American Chemical Society: Washington,  DC, 1986.

3.    1985 Annual Book of ASTM Standards, Vol. 11.01; "Standard Specification for
Reagent Water"; ASTM: Philadelphia,  PA, 1985; D1193-77.
                                   7061A  -  5                       Revision 1
                                                                  July 1992

-------
                                      METHOD  7061A
                 ARSENIC  (ATOMIC ABSORPTION, GASEOUS HYDRIDE)
7.1 Turn off
 heat,cool,
and add HNO.
7.3 Tr«n»fer
standard* to
 flask*,add
•ample,bring
  to volume
                    C   Start    J
                       7.1 Place
                      aliquot of
                       dige*ted
                       •ample in
                        beaker
                        7.1 Add
                     concentrated
                    HNO, and H.SO.;
                       evaporate
                        •ample
                     7.1 Continue
                      adding HNO.
                      to complete
                       digestion
                       7.1 Cool
                      •ample,add
                     reagent H,0,
                    evaporate,cool
                     7.1 Transfer
                    digested *ample
                     to flaik.add
                    cone HC1,bring
                       to volume
7 . 2 Prepare
stand*
irds .
transfer to
flasks
bring
to volume


                        7 4 Add
                    prepared samplt
                    to flask,bring
                     to volume,use
                       a* blank
 7.S Transfer
  portion of
digested sample
or standard  to
reaction vessel
  7.5 Add  KI
 solution,  and
     SnCl,
   solution
  7.5 Reduce
 metal to  its
    lowest
   oxidation
     state
  7.5 Attach
 vessel  to  gas
glassware,fill
dropper  with  Zn
    s1urry
 7.S Introduce
   Zn slurry
  into sample
  or standard
   solution
    7 6 Use
193.7-nm wave-
  length and
  background
  correction
                                        7.7 To operate
                                        argon hydrogen
                                          flame,follow
                                          manufacturer
                                          instructions
                     method of
                     •tandard
                     addition
                       u*ed?
                   7.8 Plot
                abaorbanee* of
                •piked (ample*
                   blank v*.
                concentration*
                     7.8 Have
                   concentration
                    be part of
                    calibration
                  c
Stop
                                                 7061A -  6
                                                   Revision  1
                                                   July  1992

-------
                                 METHOD 7062

       ANTIMONY AND ARSENIC  (ATOMIC ABSORPTION,  BOROHYDRIDE REDUCTION)


1.0  SCOPE AND APPLICATION

      1.1  Method 7062 is an atomic absorption procedure for determining  1 jjg/l
to 400//g/L concentrations of antimony and arsenic in wastes, mobility procedure
extracts, soils, and ground water.   Method 7062 is  approved for sample matrices
that contain up to  a  total of 4000 mg/L concentrations of cobalt, copper, iron,
mercury,  or  nickel.  A  solid  sample can  contain  up  to 40% by  weight of the
interferents  before exceeding 4000  mg/L  in  a digested sample.   All  samples
including aqueous matrices must be  subjected to an  appropriate dissolution step
prior to analysis.  Spiked samples and relevant standard reference materials are
used to determine the applicability of the method to a given waste.

2.0  SUMMARY OF METHOD

      2.1  Samples are prepared according to the nitric acid digestion procedure
described   in   Method   3010  for   aqueous   and  extract  samples   and  the
nitric/peroxide/hydrochloric acid digestion procedure described in Method 3050
(furnace  AA  option)  for  sediments,  soils,  and  sludges.  Excess  peroxide is
removed  by  evaporating  samples  to  near dryness at  the end of  the digestion
followed  by  degassing the  samples  upon addition of urea.   L-cysteine is then
added as  a  masking agent.   Next,  the  antimony and arsenic in  the digest are
reduced to the trivalent forms with  potassium iodide.  The trivalent  antimony and
arsenic are then converted to volatile hydrides using  hydrogen produced  from the
reaction  of  the  acidified sample with  sodium borohydride in  a continuous-flow
hydride generator.

      2.2  The  volatile hydrides are  swept into,  and decompose in,  a heated
quartz   cell   located   in   the   optical   path   of   an  atomic   absorption
spectrophotometer.    The  resulting  absorption  of  the  lamp  radiation  is
proportional to the arsenic  or antimony concentration.

      2.3  The typical detection limit for this method is l.Opg/L.

3.0  INTERFERENCES

      3.1   Very  high (>4000  mg/L)  concentrations of cobalt,  copper,  iron,
mercury, and nickel  can  cause analytical interferences through precipitation as
reduced metals and associated blockage of transfer lines and fittings.

      3.2  Traces of peroxides left following the sample work-up can result in
analytical interferences.  Peroxides  must  be removed by evaporating each sample
to near dryness followed by  reaction  with  urea and allowing sufficient time for.
degassing before analysis (see Sections 7.1 and 7.2).
                                    7062-1                        Revision 0
                                                                  September 1994

-------
      3.3   Even after  acid  digestion,  organic  compounds  will  remain  in the
sample.   These  flame  gases  and  these organic  compounds  can  absorb  at the
analytical wavelengths and background correction must be used.

4.0  APPARATUS AND MATERIALS

      4.1  Electric  hot  plate:   Large enough to hold at  least  several  100 mL
Pyrex digestion beakers.

      4.2   A  continuous-flow hydride  generator:    A  commercially  available
continuous-flow  sodium  borohydride/HCl  hydride  generator  or  a  generator
constructed similarly to that shown in Figure 1 (P. S.  Analytical or equivalent).

            4.2.1  Peristaltic Pump:  A four-channel,  variable-speed peristaltic
      pump to permit regulation of liquid-stream flow rates (Ismatec Reglo-100
      or equivalent).   Pump  speed and tubing diameters should  be  adjusted to
      provide  the  following  flow rates:    sample/blank  flow  =  4.2  mL/min;
      borohydride flow = 2.1 mL/min;  and potassium iodide  flow = 0.5 mL/min.

            4.2.2  Sampling Valve  (optional):     A sampling valve (found in the
      P. S.  Analytical  Hydride Generation System  or equivalent)  that  allows
      switching between  samples and blanks  (rinse  solution) without introduction
      of air into the system will  provide  more signal stability.

            4.2.3  Transfer Tubing and Connectors: Transfer tubing  (1 mm I.D.),
      mixing T's, and connectors are made of a fluorocarbon  (PFA  or TFM) and are
      of  compatible  sizes  to form  tight,  leak-proof  connections  (Latchat,
      Technicon, etc. flow injection  apparatus accessories or equivalent).

            4.2.4  Mixing  Coil:  A 20-turn  coil made by wrapping transfer tubing
      around a 1-cm diameter by 5-cm  long  plastic or glass rod (see Figure 1).

            4.2.5  Mixing Coil  Heater, if appropriate:  A  250-mL  Erlenmeyer
      flask containing 100 mL  of  water  heated to boiling  on  a  dedicated  one-
      beaker hotplate (Corning PC-35  or equivalent).   The  mixing coil in 4.2.4
      is immersed in  the boiling water to speed kinetics of the hydride forming
      reactions  and  increase  solubility   of   interfering   reduced   metal
      precipitates.

            4.2.6  Gas-Liquid Separator:  A glass apparatus for collecting and
      separating liquid  and  gaseous  products  (P.T.  Analytical accessory or
      equivalent) which  allows the liquid fraction to  drain  to waste and gaseous
      products above the liquid to be swept by a regulated carrier gas (argon)
      out of the cell for  analysis.   To avoid undue  carrier gas dilution, the
      gas volume above the liquid  should not  exceed 20 mL.  See Figure 1 for an
      acceptable separator shape.

            4.2.7  Condenser:  Moisture picked  up by the  carrier  gas  must be
      removed before encountering the hot  absorbance cell.   The moist carrier
      gas with the  hydrides is dried by passing the gasses  through a small  (< 25

                                    7062-2                        Revision 0
                                                                  September 1994

-------
      ml) volume condenser coil (Ace Glass Model  6020-02 or equivalent) that is
      cooled to 5°C by a water chiller (Neslab RTE-110 or equivalent).  Cool tap-
      water in place of a chiller is acceptable.

            4.2.8  Flow Meter/Regulator:  A meter capable  of  regulating up to 1
      L/min of argon carrier gas is recommended.

      4.3  Absorbance Cell:   A 17 cm or longer quartz tube  T-cell  (windowless is
strongly suggested)  is recommended,  as  shown  in  Figure 1  (Varian Model  VGA-76
accessory or equivalent).   The cell  is held in  place by a  holder that positions
the cell  about 1  cm over  a conventional AA  air-acetylene  burner head.   In
operation, the cell  is heated to around 900°C.

      4.4  Atomic  absorption spectrophotometer:  Single or  dual channel, single-
or  double-beam instrument   having  a  grating  monochromator,  photomultiplier
detector, adjustable slits,  a wavelength range  of 190 to 800  nm, and provisions
for interfacing with  an appropriate recording device.

      4.5  Burner: As recommended by the particular instrument manufacturer for
an air-acetylene flame.  An  appropriate mounting  bracket attached to the burner
that suspends the  quartz absorbance cell between 1 and 2 cm above the burner slot
is required.

      4.6   Antimony  and  arsenic hollow cathode  lamps  or  antimony  and arsenic
electrodeless discharge lamps  and power supply.   Super-charged hollow-cathode
lamps or EDL lamps are recommended for maximum sensitivity.

      4.7    Strip-chart  recorder  (optional):        Connect  to  output  of
spectrophotometer.

5.0  REAGENTS

      5.1   Reagent water:   Water must be monitored  for impurities.   Refer to
Chapter 1 for definition of Reagent water.

      5.2  Concentrated nitric acid  (HN03):  Acid must be  analyzed to determine
levels of impurities.  If a method blank is 
-------
                                  QUARTZ CELL

                                  A A  BURNER
                                                                        TO
                                                                     CHILLER
     DISCONNECTED)
    OURI NO  S*/8n
   ...  ftMALYJIS  J
                                                                OM/L1QUIO
                                                                SEPARATOR
                                                                 _—» DRAIN
                         20 TORN COIL
                           (TEFLON)
                           HOTPLATE-
                           MALWt
                          (•LANK)
Figure 1. Continuous-flow sodium borohydride/hydride generator apparatus set-up
and an AAS sample  introduction  system.
                                     7062-4
Revision 0
September 1994

-------
      5.6  Urea (H2NCONH2):  A 5.00-g portion of reagent grade urea must be  added
to  a 25-mL  aliquot of  each sample  for removal  of excess  peroxide through
degassing  (see Section 7.2).

      5.7  L-cysteine (C6H12N204S2):  A 1.00-g portion of reagent grade  L-cystine
must  be  added to a  25-mL aliquot  of  each  sample for  masking the effects of
suppressing transition metals (see  Section 7.2).

      5.8  20% Potassium iodide (KI):  A 20% KI  solution (20 g  reagent-grade KI
dissolved  and  brought to  volume  in  100 ml reagent water) must  be  prepared for
reduction  of antimony and arsenic to their +3 valence states.

      5.9  4% Sodium borohydride (NaBH4):  A 4%  sodium borohydride  solution (20
g reagent-grade NaBH4 plus 2 g sodium hydroxide dissolved in 500 ml of reagent
water) must  be prepared for  conversion  of  the antimony  and  arsenic  to  their
hydrides.

      5.10 Analyte solutions:

            5.10.1  Antimony and arsenic stock standard solution  (1,000 mg/L):
      Either procure certified aqueous standards from a supplier and  verify by
      comparison  with  a  second  standard,  or  dissolve  1.197 g  of  antimony
      trioxide Sb203 and  1.320 g  of  arsenic  trioxide As203 in 100 ml of reagent
      water containing 4 g NaOH.  Acidify the solution with 20  ml  concentrated
      HN03 and dilute to 1 liter.

            5.10.2   Intermediate antimony  and  arsenic solution:   Pipet  1 ml
      stock antimony and  arsenic solution into a  100-mL  volumetric flask and
      bring  to  volume  with  reagent  water  containing  1.5   ml   concentrated
      HN03/liter (1  mL  = 10 fjg each  of Sb and  As).

            5.10.3   Standard  antimony  and arsenic  solution:   Pipet  10 mL
      intermediate antimony and  arsenic solution into a  100-mL  volumetric  flask
      and  bring  to  volume with  reagent  water containing 1.5  mL  concentrated
      HN03/liter (1  mL  = 1 jjg each of Sb and As).

6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   All samples must  have been  collected  using a sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2  All sample containers must be prewashed with detergents, acids, and
reagent water.  Plastic and glass containers are both suitable.

      6.3   Special   containers   (e.g.,  containers  used  for volatile organic
analysis) may have to be used if very volatile antimony and arsenic compounds are
suspected to be present in the samples.

      6.4  Aqueous samples must be acidified to a pH of <2 with nitric acid.


                                    7062-5                        Revision 0
                                                                  September 1994

-------
      6.5  Nonaqueous samples shall be refrigerated,  when  possible, and analyzed
as soon as possible.

7.0  PROCEDURE

      7.1  Place a 100-mL portion of an  aqueous  sample or extract or 1.000 g of
a dried solid sample in a 250-mL digestion beaker.  Digest aqueous samples and
extracts according to Method 3010.  Digest solid samples according to Method 3050
(furnace AA option) with the  following modifications:  add 5 ml of concentrated
hydrochloric  acid  just prior  to  the final volume  reduction  stage to  aid in
antimony recovery; the final volume reduction should  be to less than 5 ml but not
to dryness  to adequately remove  excess  hydrogen  peroxide (see  note).   After
dilution to volume, further dilution with diluent may be necessary if analytes
are known to exceed 400 fjg/l or if interferents are expected to exceed 4000 mg/L
in the digestate.

            Note:  For solid  digestions,  the volume  reduction stage is critical
            to obtain accurate data, especially for arsenic.   Close monitoring
            of each sample is necessary when this critical stage is reached.

      7.2  Prepare samples for  hydride analysis by adding 5.00 g  urea, 1.00 g L-
cysteine, and 20 mL concentrated HC1  to  a 25-mL  aliquot of digested sample in a
50-mL volumetric flask.  Heat in a water bath until the L-cysteine has dissolved
and  effervescence  has  subsided   (At  least  30  minutes  is  suggested.    If
effervescense is still  seen, repeat step  7.1 with more volume reduction.).  Bring
flask to volume with reagent  water before analyzing.  A 1:1 dilution correction
must be made in the final concentration calculations.

      7.3   Prepare working standards from the standard  antimony  and  arsenic
solution.   Transfer 0, 0.5,  1.0,  1.5,  2.0,  and 2.5 ml  of standard  to 100-mL
volumetric flasks and bring to volume with diluent.   These concentrations will
be 0, 5, 10, 15, 20, and 25 //g Sb and As/liter.

      7.4   If EP extracts (Method  1310)  are  being analyzed for arsenic,  the
method  of standard  additions  must  be  used.   Spike  appropriate amounts  of
intermediate or standard antimony and arsenic solution to three 25 ml aliquots
of each  unknown.   Spiking volumes should be kept less than 0.250  ml  to avoid
excessive spiking dilution errors.

      7.5   Set up  instrumentation  and  hydride generation apparatus  and fill
reagent containers.  The sample and blank flows should be  set around 4.2 mL/min,
the borohydride flow around 2.1 mL/min, and the potassium  iodide  flow around 0.5
mL/min.  The argon  carrier gas flow is adjusted to about 200 mL/min.  For the AA,
use the 217.6-nm wavelength and 0.7-nm slit  width (or manufacturer's recommended
slit-width) without  background correction  if  analyzing  for antimony.   Use the
193.7-nm wavelength and 0.7-nm slit width (or manufacturer's  recommended slit-
width) with background correction for the analysis of arsenic.  Begin all flows
and allow 10 minutes for warm-up.
                                    7062-6                        Revision 0
                                                                  September 1994

-------
      7.6  Place sample feed line into a prepared sample solution  and start pump
to begin  hydride generation.   Wait  for  a maximum steady-state  signal  on the
strip-chart recorder  or output meter.   Switch  to blank sample  and  watch for
signal to decline to baseline  before  switching to  the next sample and beginning
the next analysis.  Run standards first (low to high), then unknowns.  Include
appropriate QA/QC solutions, as required.   Prepare calibration curves and convert
absorbances to concentration.    If a heating coil  is  not being  used, KI must be
added to the samples and heated for thirty minutes to ensure reduction.

        CAUTION:  The  hydrides  of  antimony  and  arsenic  are  very  toxic.
                  Precautions must be taken to avoid inhaling the gas.

      7.7  If the method of standard additions was employed, plot the measured
concentration  of the  spiked  samples  and unspiked  sample  versus the  spiked
concentrations.  The spiked concentration axis intercept will be the method of
standard additions  concentration.   If the plot does not result  in a straight
line, a  nonlinear  interference  is present.   This  problem can  sometimes  be
overcome by dilution or addition of  other reagents  if there is some knowledge
about the waste.   If the method of standard additions  was not required, then the
concentration is determined from a standard calibration curve.

8.0  QUALITY CONTROL

      8.1  See section 8.0 of Method 7000.

9.0  METHOD PERFORMANCE

      9.1  The relative standard deviations obtained by  a single laboratory for
7 replicates of a contaminated soil  were 18% for antimony at 9.1  ug/L in solution
and 4.6% for arsenic at 68 ug/L in solution.  The average percent recovery of the
analysis of an 8/yg/L  spike on ten different samples is  103.7% for arsenic and
95.6% for antimony.

10.0  REFERENCES

1.    Methods   for  Chemical  Analysis  of  Water  and Wastes,  EPA-600/4-82-055,
      December 1982, Method 206.3.

2.    "Evaluation of Hydride  Atomic Absorption  Methods  for  Antimony,  Arsenic,
      Selenium, and  Tin",  an EMSL-LV internal report under Contract 68-03-3249,
      Job Order  70.16,  prepared  for T.  A.  Hinners by  D.  E.  Dobb, and  J.  D.
      Lindner  of Lockheed Engineering and Sciences Co., and  L.  V. Beach of the
      Varian Corporation.
                                    7062-7                        Revision 0
                                                                  September 1994

-------
                                    METHOD  7062
ANTIMONY  AND ARSENIC  (ATOMIC ABSORPTION,  BOROHYDRIDE REDUCTION)
     7.1 Use Method
    3050 (furnace AA
    option) to digest
      1.0 g sample.
         7.1 Add
       concentrated
          HCI.
       7.1 Do final
          volume
       reduction and
        dilution, as
        described
            Yes
        7.1 Further
        dilute with
         diluent.
  7.2 Add to
  aliquot urea:
L-cysteine, HCI:
heat H20 bath;
bring to volume.
  7.3 Prepare
 standards from
 standard stock
 solutions ot Sb
    and As.
7.6 - 7.6 Analyze
   the sample
  using hydride
   generation
   apparatus.
                                  7.6 • 7.7 Determine
                                   Sb and As cone.
                                     from standard
                                      calibration
                                        curve.
                                 7.1 Use
                               Method 3010
                               to digest  100
                                ml sample.
                                                                  7.4 Use the
                                                                   method of
                                                                   standard
                                                                 additions on EP
                                                                 extracts, only.
7.6 -7.6 Analyze
   the sample
 using hydride
   generation
   apparatus.
                               7.7 Determine
                                 Sb and As
                               concentrations
                               by Method of
                             Standard Additions.
                                       Stop
                                        7062-8
                                               Revision 0
                                               September 1994

-------
                                 METHOD 7080A

                 BARIUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

      1.1   See Section 1.0 of Method 7000.

2.0  SUMMARY OF METHOD

      2.1   See Section 2.0 of Method 7000.

3.0  INTERFERENCES

      3.1   See Section 3.0 of Method 7000 if interferences are suspected.

      3.2   High hollow cathode current settings and a narrow spectral band pass
must  be used,  because  both   barium  and  calcium emit  strongly  at  barium's
analytical wavelength.

      3.3   Barium  undergoes   significant  ionization  in  the  nitrous  oxide/
acetylene flame, resulting in a significant decrease in sensitivity.  All
samples and standards must contain a ionization suppressant.  The type of
suppressant and concentration used must be documented.

4.0  APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument parameters (general):

            4.2.1  Barium hollow cathode lamp.
            4.2.2  Wavelength:  553.6 nm.
            4.2.3  Fuel:  Acetylene.
            4.2.4  Oxidant:  Nitrous oxide.
            4.2.5  Type of flame:  Fuel  rich.
            4.2.6  Background correction:  Not required.

5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards:

            5.2.1    Stock  solution:    Dissolve  1.7787  g  barium  chloride
      (BaCl22H20)  analytical  reagent grade  in reagent water and  dilute to 1
      liter (1000 mg/L).   Alternatively,  procure a certified standard  from a
      supplier and verify by comparison  with a second standard.

            5.2.2    Prepare  dilutions  of the  stock solution to  be used  as
      calibration standards at the time of analysis.  The calibration standards


                                  7080A  -  1                       Revision 1
                                                                  September 1994

-------
      should  be  prepared  using  the  same  type  of  acid  and   at  the  same
      concentration  as  will  result  in  the   sample  to  be  analyzed  after
      processing.  All   calibration   standards  and samples should contain the
      ionization suppressant.

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See Chapter Three,  Section  3.1.3, Sample Handling and Preservation.

7.0   PROCEDURE

      7.1   Sample preparation: The procedures for  preparation of the  sample are
given in Chapter Three, Section 3.2.

      7.2   See Method 7000, Section 7.2, Direct Aspiration.

8.0  QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.

9.0  METHOD PERFORMANCE

      9.1   The performance characteristics for an aqueous sample  free of inter-
ferences are:

      Optimum concentration range:  1-20 mg/L with a wavelength of 553.6 nm.
      Sensitivity:  0.4 mg/L.
      Detection limit:  0.1 mg/L.

      9.2   In a single laboratory, analysis of a mixed industrial-domestic waste
effluent, digested with Method  3010, at concentrations of 0.4 and  2 mg Ba/L gave
standard deviations  of +0.043 and ±0.13,  respectively.    Recoveries at these
levels were 94% and 113%, respectively.

10.0  REFERENCES

1.    Methods  for  Chemical  Analysis  of Water  and Wastes,  EPA-600/4-82-055,
December 1982, Method 208.1.
                                   7080A  -  2                       Revision 1
                                                                  September 1994

-------
                   METHOD 7080A
BARIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)
            (    Start     J
               5.2 Prepare
               standards.
              7.1 For sample
              preparation see
            Chapter 3, Section
                  3.2.
             7.2 Analyze using
               Method 7000
                Section 7.2.
             f     Stop    J
                      7080A  -  3
Revision  1
September 1994

-------
                                 METHOD  7081
                 BARIUM  (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)

1.0  SCOPE AND APPLICATION
      1.1   See Section 1.0 of Method 7000.

2.0  SUMMARY OF METHOD
      2.1   See Section 2.0 of Method 7000.

3.0  INTERFERENCES
      3.1   See Section 3.0 of Method 7000.
      3.2   Barium is known to form  a  barium  carbide  in the graphite furnace.
This less volatile carbide can cause losses  of sensitivity and memory effects.
      3.3   The long residence time and the high concentration of the analyte in
the optical  path  of the graphite furnace can lead to severe physical and chemical
interferences.  Furnace parameters must be optimized to minimize these effects.
      3.4   Because of possible chemical interaction, nitrogen should not be used
as a purge gas.
      3.5   Halide acids should not be used.

4.0  APPARATUS AND MATERIALS
      4.1   For basic apparatus, see Section 4.0 of Method 7000.
      4.2   Instrument parameters (general):
            4.2.1   Drying time and temp:  30 sec at 125°C.
            4.2.2   Ashing time and temp:  30 sec at 1200°C.
            4.2.3   Atomizing time and temp:  10 sec at 2800°C.
            4.2.4   Purge gas:  Argon (nitrogen should not be used).
            4.2.5   Wavelength:   553.6 nm.
            4.2.6   Background correction:   Not required.
                                   7081 - 1                       Revision 0
                                                                  July 1992

-------
            4.2.7   Other operating  parameters should be set as specified by the
      particular instrument manufacturer.

      NOTE: The above concentration  values and instrument conditions are for a
            Perkin-Elmer  HGA-2100,  based  on  the  use  of  a 20-uL  injection,
            continuous-flow purge gas,  and  nonpyrolytic graphite.  Smaller size
            furnace devices or those employing faster rates of atomization can
            be operated using lower atomization  temperatures  for  shorter time
            periods than the above-recommended settings.


5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards

            5.2.1   Stock solution - Dissolve 1.7787 g barium chloride
      (BaCl2 2H20,  analytical reagent  grade)  in  water and dilute  to  1  liter.
      Alternatively, procure a certified  standard from a supplier and verify by
      comparison with a second standard.

            5.2.2   Prepare  dilutions  of  the  stock  solution  to  be used  as
      calibration standards at the time of  analysis.  The calibration standards
      should  be  prepared  using the  same  type  of  acid and  at  the  same
      concentrations as in the sample after processing (0.5% v/v HN03).


6.0  SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See Chapter Three, Step 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

      7.1   Sample Preparation - The procedures for preparation of the sample are
given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.3, Furnace Technique.


8.0  QUALITY ASSURANCE

      8.1   See Section 8.0 of Method 7000.
9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are not available at this time.
                                   7081 - 2                       Revision 0
                                                                  July 1992

-------
10.0  REFERENCES

1.    Methods for  Chemical  Analysis of  Mater and Wastes;  U.S.  Environmental
Protection Agency.  Office of Research and Development. Environmental Monitoring
and Support  Laboratory.  ORD Publication  Offices  of Center  for  Environmental
Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.
                                   7081 - 3                       Revision 0
                                                                  July 1992

-------
                    METHOD  7081
BARIUM (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)
                       Start
                     5.0 Pr«par«
                      standard*
                   7 1 For aampla
                   preparation »••
                  Chaptar 3, Saetion
                        3.2
                  7 2 Analyza uung
                    M.lhod 700C
                     Section 7 3
                       Stop
                        7081  -  4
Revision 0
July  1992

-------
                                 METHOD 7090

              BERYLLIUM (ATOMIC ABSORPTION, DIRECT  ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction may be required because nonspecific absorption
and light scattering can be significant at the analytical wavelength.

     3.3  Concentrations  of  aluminum  greater  than  500  ppm  may  suppress
beryllium absorbance.  The addition of  0.1% fluoride has been found effective
in eliminating  this  interference.    High  concentrations  of  magnesium and
silicon cause similar problems and require  the  use of the method of standard
additions.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Beryllium hollow cathode lamp.
          4.2.2  Wavelength:   234.9 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Nitrous oxide.
          4.2.5  Type of  flame:  Fuel  rich.
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section  5.0 of  Method 7000.

     5.2   Preparation  of  standards;

           5.2.1  Stock solution:  Dissolve 11.6586 g beryllium  sulfate, BeSO^
     in Type II water   containing  2   mL  nitric   acid   and  dilute  to  1 liter.
                                   7090 - 1
                                                          Revision
                                                          Date  September  1986

-------
     Beryllium metal  can also be dissolved 1n  H2S04.   Alternatively, procure  a
     certified standard from a supplier and verify  by comparison with  a  second
     standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution   to   be  used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  add and   at   the same
     concentration  as  will  result  1n  the  sample  to  be   analyzed   after
     processing (0.5% v/v HMOs).


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample  Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample Preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000,  Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics for an aqueous sample free of Inter-
ferences are:

     Optimum concentration  range:  0.05-2 mg/L with a wavelength of 234.9 nm.
     Sensitivity:  0.025 mg/L.
     Detection  limit:  0.005 mg/L.

     9.2  In  a  single  laboratory,  analysis  of  a mixed industrial-domestic
waste effluent, digested with Method 3010,  at concentrations of 0.01 and 0.25
mg/L gave standard deviations of   +0.001  and +0.002, respectively.  Recoveries
at these levels were  100% and 97%, respectively.

     9.3  For concentrations of beryllium below  0.02 mg/L, the furnace proce-
dure  (Method  7091) is  recommended.


10.0  REFERENCES

1.   Methods  for   Chemical   Analysis   of  Water  and Wastes, EPA-600/4-82-055,
December  1982,  Method 210.1.
                                   7090 - 2
                                                          Revision
                                                          Date   September 1986

-------
                   METHOD 709O

BERYLLIUM (ATOMIC  ABSORPTION. DIRECT ASPIRATION)
               C
                 5.0
                    Prepare
                    standards
                 7.1
               	1 For
                     •ample
               preparation see
                   chapter 3.
                  section 3.2
                7.Z
                Analyze ualng
                 Method 700O.
                 Section 7.2
              f     Stop      J
             7090 - 3
                                       Revision       0
                                       Date   September 1986

-------
                                 METHOD 7091

              BERYLLIUM (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  The long residence  time  and  high  concentrations  of the atomized
sample in the  optical  path  of  the  graphite  furnace  can result 1n severe
physical and chemical interferences.   Furnace parameters must be optimized to
minimize these effects.

     3.3  In addition to the  normal interferences experienced during graphite
furnace analysis, beryllium analysis   can  suffer  from severe nonspecific ab-
sorption and light scattering caused   by matrix components during atomlzation.
Simultaneous background  correction  is  required  to  avoid  erroneously high
results.


4.0  APPARATUS AND MATERIALS

     4.1  For basic  apparatus,  see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Drying time and temp:  30 sec at 125*C.
          4.2.2  Ashing time and temp:  30 sec at 1000*C.
          4.2.3  Atomizing time and temp:  10 sec at 2800*C.
          4.2.4  Purge gas:  Argon.
          4.2.5  Wavelength:  234.9 nm.
          4.2.6  Background correction:  Required.
          4.2.7  Other operating parameters should be  set as specified by the
     particular  instrument manufacturer.
     NOTE:  The  above concentration values and instrument conditions are for a
            Perkin-Elmer HGA-2100, based  on  the  use  of  a 20-uL Injection,
            continuous-flow purge  gas,  and  nonpyrolytic  graphite.  Smaller
            sizes of furnace   devices or  those  employing  faster  rates of
            atomization can be  operated  using lower atomlzation temperatures
            for  shorter time periods than the above-recommended settings.
                                   7091 -  1
                                                         Revision
                                                         Date  September 1986

-------
5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:  Dissolve 11.6586 g beryllium sulfate, BeS04,
     In Type II water containing 2 ml concentrated nitric acid and dilute to
     1 liter.  Beryllium metal can  also be dissolved in acid.  Alternatively,
     procure a certified standard  from  a  supplier  and verify by comparison
     with a second standard.

          5.2.2  Prepare dilutions of the stock  solution  to be used as cali-
     bration standards at the  time  of  analysis.   The calibration standards
     should  be  prepared  using  the  same  type  of  acid  and  at  the same
     concentrations as in the sample after processing (0.5% v/v HN03).


6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

     6.1  See  Chapter Three, Section 3.1.3, Sample Handling and Preservation.

7.0  PROCEDURE

     7.1  Sample  Preparation;   The  procedures  for  preparation of the  sample
are given in Chapter Three,  Section 3.2.

     7.2  See  Method 7000,  Paragraph 7.3,  Furnace  Procedure.  The calculation
is given  in Method 7000,  Paragraph 7.4.


8.0  QUALITY CONTROL

     8.1  See  Section 8.0 of Method 7000.

9.0  METHOD  PERFORMANCE

     9.1  Precision and  accuracy data  are  not available  at this time.

     9.2  The  performance characteristics  for an  aqueous sample free  of  inter-
ferences  are;

     Optimum concentration  range: 1-30  ug/L.
     Detection limit:  0.2  ug/L.


10.0   REFERENCES

1.  Methods for   Chemical   Analysis   of  Water  and Wastes,  EPA-600/4-82-055,
December  1982, Method  210.2.


                                   7091 - 2
                                                          Revision       0
                                                          Date  September 1986

-------
                  METHOD 7091

BERYLLIUM (ATOMIC ABSORPTION.  FURNACE TECHNIQUE)
                      Prepare
                     et*nderds
7.1
prepar
cl-
•ec
For '
•ample
•atlon aee
i»pter 3.
tlon 3.2
                  7.Z
                  Analyze using
                   Method 7000.
                   Section 7.3.
                 calculation 7.4
               f     Stop      J
            7091 - 3
                                      Revision       0
                                      Date   September 1986

-------
                                 METHOD 7130

               CADMIUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 1f Interferences are suspected.

     3.2  Nonspecific absorption and  light  scattering  can be significant at
the analytical wavelength.  Thus background correction is required.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Cadmium hollow  cathode  lamp.
          4.2.2  Wavelength:  228.8 nm.
          4.2.3  Fuel:   Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:  Dissolve  1.000  g cadmium metal  (analytical
     reagent grade)  in 20 mL of  1:1  HNOs  and  dilute to 1 liter with  Type II
     water.  Alternatively, procure a  certified  standard from a supplier and
     verify by comparison with a second  standard.

          5.2.2  Prepare dilutions of the stock  solution  to be used as cali-
     bration standards at the  time  of  analysis.   The calibration standards
     should  be  prepared  using the  same  type  of  add  and  at  the same
                                  7130 - 1
                                                         Revision
                                                         Date  September 1986

-------
     concentration  as  will  result  1n  the  sample  to  be  analyzed  after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics  for an aqueous sample free of Inter-
ferences are:

     Optimum concentration  range: 0.05-2 mg/L with a wavelength of 228.8 nm.
     Sensitivity:  0.025 mg/L.
     Detection  Hm1t:  0.005 mg/L.

     9.2  For concentrations of cadmium below 0.02 mg/L, the furnace procedure
 (Method 7131) 1s  recommended.

     9.3  Precision  and accuracy  data  are  available  1n Method 213.1 of Methods
 for Chemical Analysis  of  Water and  Wastes.

     9.4  The data shown  1n Table  1   were  obtained from  records of state  and
 contractor  laboratories.   The data  are  Intended  to show  the precision of  the
 combined  sample preparation and analysis method.


 10.0  REFERENCES

 1.   Methods  for   Chemical   Analysis  of  Water  and Wastes, EPA-600/4-82-055,
 December  1982,  Method 213.1.

 2.   Gasklll, A., Compilation and Evaluation  of  RCRA  Method Performance  Data,
 Work Assignment No.  2, EPA Contract No. 68-01-7075,  September  1986.
                                   7130 - 2
                                                          Revision
                                                          Date   September 1986

-------
                          TABLE 1. METHOD PERFORMANCE DATA
       Sample                      Preparation                Laboratory
       Matrix                        Method                   Replicates


Emission control dust                 3050                  2,770, 1,590 ug/g

Wastewater treatment sludge           3050                 12,000, 13,000 ug/g
                                   7130 - 3
                                                          Revision
                                                         Date  September  1986

-------
                 METHOD 7130

CADMIUM (ATOMIC ABSORPTION. DIRECT ASPIRATION)
9.0

Prepare
atandards


7.1 1
1 For
•ample
preparation see
chapter 3.
section 3.2


7.2 {
Analyze using
Method 7000.
Section 7.2
              f     Stop      J
           7130 - 4
                                    Revision       0
                                    Date  September 1986

-------
                                 METHOD 7131A

                CADMIUM  (ATOMIC ABSORPTION, FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

      1.1   See Section  1.0 of Method 7000.


2.0  SUMMARY OF METHOD

      2.1   See Section  2.0 of Method 7000.


3.0  INTERFERENCES

      3.1   See Section  3.0 of Method 7000 if interferences are suspected.

      3.2   In addition to the normal  interferences experienced during graphite
furnace analysis,  cadmium analysis can suffer from severe nonspecific absorption
and light scattering caused by matrix components during atomization. Simultaneous
background correction is required to avoid erroneously high results.

      3.3   Excess  chloride  may  cause premature  volatilization  of  cadmium.
Ammonium  phosphate used  as  a  matrix modifier  minimizes  this  loss.   Other
modifiers may be used as long as it is documented with the type of suppressant
and concentration.

      3.4   Many plastic pipet  tips  (yellow)  contain cadmium.   Use "cadmium-
free" tips.


4.0  APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument parameters (general):

            4.2.1    Drying time and temp:   30 sec at 125°C.

            4.2.2    Ashing time and temp:   30 sec at 500°C.

            4.2.3    Atomizing time and temp:  10 sec at 1900°C.

            4.2.4    Purge gas:  Argon.

            4.2.5    Wavelength:   228.8 nm.

            4.2.6    Background correction:  Required.

            4.2.7   Other operating parameters should be  set  as specified by the
      particular instrument manufacturer.


                                  7131A -  1                       Revision 1
                                                                  September 1994

-------
            NOTE: The above concentration values and instrument conditions are
            for a Perkin-Elmer HGA-2100,  based on the use of a 20-uL injection,
            continuous-flow purge gas, and nonpyrolytic graphite.  Smaller sizes
            of furnace devices or  those  employing  faster  rates of atomization
            can be  operated  using lower atomization temperatures  for shorter
            time periods than the above-recommended settings.


5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards:

            5.2.1     Stock  solution:   Dissolve  1.000  g of cadmium  metal
      (analytical reagent grade)  in 20 ml of 1:1 HN03 and dilute to  1 liter with
      reagent  water.    Alternatively,  procure a  certified  standard from  a
      supplier and verify by comparison with a second standard.

            5.2.2   Prepare dilutions of the stock cadmium solution to be used
      as calibration  standards  at the  time  of analysis.   To each  100  ml of
      standard and sample alike add 2.0 ml of the ammonium phosphate solution.
      The calibration standards should be prepared to contain 0.5% (v/v)  HN03.

            5.2.3     Ammonium phosphate  solution   (40%):   Dissolve  40  g of
      ammonium phosphate,  (NH4)2HP04 (analytical reagent grade), in reagent water
      and dilute to 100 ml.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See Chapter Three, Section 3.1.3,  Sample Handling and Preservation.


7.0  PROCEDURE

      7.1   Sample preparation:  The procedures for preparation of the sample are
provided in Chapter Three, Section 3.2.

      7.2   See Method 7000,  Section 7.3, Furnace Procedure. The calculation is
provided in Method 7000, Section 7.4.


8.0  QUALITY CONTROL

      8.1   Refer to Section 8.0 of Method 7000 .


9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are available in Method 213.2 of Methods
for Chemical Analysis of Water and Wastes.
                                   7131A -  2                       Revision 1
                                                                  September 1994

-------
      9.2   The performance characteristics for an aqueous sample free of  inter-
ferences are:

      Optimum concentration range:  0.5-10 ug/L.
      Detection limit:  0.1 ug/L.

      9.3   The data  shown  in  Table  1  were obtained from records of  state and
contractor laboratories.   The  data are intended to  show the precision of the
combined sample preparation and analysis method.


10.0  REFERENCES

1.    Methods  for Chemical Analysis  of  Water  and  Wastes,  EPA-600/4-82-055,
December 1982, Method 213.2.

2.    Gaskill, A., Compilation and Evaluation of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075, September 1986.
                                  7131A - 3                       Revision 1
                                                                  September 1994

-------
                       TABLE 1. METHOD PERFORMANCE  DATA
Sample                      Preparation              Laboratory
Matrix                        Method                 Replicates
Lagoon soil                     3050           0.10,  0.095  ug/g

NBS SRM 1646 Estuarine sediment 3050                   0.35  ug/ga

Solvent extract of oily waste   3030             1.39,  1.09  ug/L



aBias  of -3% from expected value.
                                   7131A - 4                       Revision 1
                                                                   September 1994

-------
                  METHOD 7131A
CADMIUM  (ATOMIC ABSORPTION, FURNACE  TECHNIQUE)
             f     Start    J
>
f
5.2 Prepare
standards.
>
r
7. 1 For sample
preparation see
Chapter 3, Section
3.2.
>
t
7.2 Analyze using
Method 7000
Section 7.3.
>
\f
              (    Stop    J
                  7131A  -  5
Revision 1
September 1994

-------
                                 METHOD 7140

               CALCIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000.

     3.2  All elements forming stable oxyanions  (P,  B,  Si, Cr, S, V, Ti, Al,
etc.) will complex calcium and interfere  unless lanthanum is added.  Addition
of lanthanum to prepared samples  rarely  presents a problem because virtually
all environmental samples contain sufficient calcium to require dilution to be
in the  linear range of the method.

     3.3  P04, 504, and Al are  interferents.    High concentrations of Mg, Na,
and K interfere.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Calcium hollow cathode  lamp.
          4.2.2  Wavelength:  422.7 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Nitrous oxide.
          4.2.5  Type of flame:  Stoichiometric.
          4.2.6  Background correction:  Not required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:  Suspend 2.500  g of CaCOa  (analytical reagent
     grade, dried for 1 hr at 180*C) 1n  Type II  water and dissolve by adding a
                                  7140 - 1
                                                         Revision
                                                         Date  September 1986

-------
     minimum  of  dilute  HC1.     Dilute   to   1   liter  with  Type   II  water.
     Alternatively,  procure a certified standard  from a supplier  and verify by
     comparison with a second standard.

          5.2.2  Prepare  dilutions   of   the   stock  solution   to  be  used as
     calibration standards at the time of analysis.  The  calibration standards
     should  be  prepared  using  the same   type of  acid  and  at the  same
     concentration  as  will   result   in   the sample  to be   analyzed after
     processing, including 1 ml  of   lanthanum chloride   per   10 ml sample or
     standard (see Paragraph 5.2.3).

          5.2.3  Lanthanum chloride  solution: Dissolve   29  g  13203 in 250 ml
     concentrated HC1 -
          CAUTION:  REACTION IS VIOLENT -
     and dilute to 500 mL with Type  II water.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3,  Sample Handling  and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section  8.0  of Method  7000.
9.0  METHOD  PERFORMANCE

     9.1   Precision and accuracy data are available in Method 215.1 of Methods
for Chemical Analysis of Water and Wastes.

     9.2   The  performance  characteristics  for  an  aqueous  sample  free of
interferences are:

     Optimum concentration  range:  0.2-7 mg/L with a wavelength of 422.7 nm.
     Sensitivity:  0.08 mg/L.
     Detection limit:  0.01 mg/L.


10.0   REFERENCES

1.   Methods for   Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December  1982, Method 215.1.


                                  7140 - 2
                                                         Revision      0
                                                         Date  September 1986

-------
                METHOD 7140

CALCIUM  (ATOMIC ABSORPTION. DIRECT ASPIRATION)
              f     Start     J

s.o


Prepare
standards

7.1
SI
preparat
chat
sect!


•or
imple
.Ion see
iter 3.
on 3.2

J.2 1
Analyze using
Method 7000.
Section 7.2


              f     Stop      J
            7140 -  3
                                     Revision       0
                                     Date  September 1986

-------
                                 METHOD 7190

               CHROMIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)
1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  An ionization interference may occur  if the samples have a signifi-
cantly higher alkali metal content  than  the standards.  If this interference
is encountered,  an  ionization  suppressant  (KC1)  should  be  added to both
samples and standards.

     3.3  Background correction may be required because nonspecific absorption
and scattering can be  significant  at  the analytical wavelength.  Background
correction with certain instruments may be difficult at this wavelength due to
low-intensity output from hydrogen or  deuterium  lamps.  Consult the specific
instrument manufacturer's literature for details.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Chromium hollow cathode lamp.
          4.2.2  Wavelength:  357.9 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Nitrous oxide.
          4.2.5  Type of flame:  Fuel rich.
          4.2.6  Background correction:  Not required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.
                                  7190 - 1
                                                         Revision
                                                         Date  September 1986

-------
     5.2  Preparation  of standards;

          5.2.1  Stock solution:   Dissolve  1.923 g of chromium trloxlde  (003,
     analytical reagent grade)   1n  Type II  water,  acidify with redistilled
     HN03, and dilute  to 1 liter.   Alternatively, procure a certified  standard
     from a supplier and verify by comparison with a second standard.

          5.2.2  Prepare dilutions of the stock  solution  to be  used  as cali-
     bration standards at the  time  of  analysis.   The calibration standards
     should  be  prepared  using  the  same type  of   add  and  at   the  same
     concentration  as  will   result  1n the  sample   to  be  analyzed  after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter  Three,  Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures   for  preparation  of the  sample
are given In Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method 7000.
9.0  METHOD PERFORMANCE

     9.1  The  performance  characteristics  for  an  aqueous  sample  free of
Interferences are;

     Optimum concentration range:  0.5-10 mg/L with a wavelength of 357.9 nm.
     Sensitivity:  0.25 mg/L.
     Detection limit:  0.05 mg/L.

     9.2  For concentrations of chromium below 0.2 mg/L, the furnace procedure
(Method 7191) 1s recommended.

     9.3  Precision and accuracy data are available 1n Method 218.1 of Methods
for  Chemical Analysis of Water and Wastes.

     9.4  The data shown in Table  1  were  obtained from records of state and
contractor  laboratories.  The data are  intended  to show the precision of the
combined  sample preparation and analysis method.
                                   7190 - 2
                                                         Revision
                                                         Date  September 1986

-------
10.0  REFERENCES
1.    Methods for Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December 1982, Method 218.1.
2.    Gasklll, A., Compilation and Evaluation of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075, September 1986.
                                  7190 - 3
                                                         Revision
                                                         Date  September 1986

-------
                          TABLE 1. METHOD PERFORMANCE DATA
       Sample                      Preparation                Laboratory
       Matrix                        Method                   Replicates


Wastewater treatment sludge          3050                    6,100, 6,000 ug/g

Emission control dust                3050                    2.0, 2.8 ug/g
                                   7190 - 4
                                                          Revision
                                                          Date  September 1986

-------
                 METHOD 719O

CHROMIUM  (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
              C
                5.O
                    Prepare
                   •tandarda
7. 1
prepar
cf
• ec
For
•ample
•atlon sea
lapter 3,
:tion 3.2

7.2


Analyz« using
Method 7000.
Section 7.2
              f     Stop      J
          7190 -  5
                                   Revision       0
                                   Date  September 1986

-------
                                 METHOD 7191

               CHROMIUM (ATOMIC ABSORPTION.  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Low  concentrations   of   calcium   and/or   phosphate   may  cause
interferences; at concentrations above 200  mg/L, calcium's effect is constant
and eliminates the effect of phosphate.  Calcium nitrate is therefore added to
ensure a  known constant effect.

     3.3  Nitrogen should not be used as  the  purge gas because of a possible
CN band interference.

     3.4  Background correction may be required because nonspecific absorption
and scattering can be  significant  at  the analytical wavelength.  Background
correction with certain instruments may be difficult at this wavelength due to
low-intensity output from hydrogen or  deuterium  lamps.  Consult the specific
instrument manufacturer's literature for details.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus,  see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Drying time and temp:  30 sec at 125*C.
          4.2.2  Ashing time and temp:  30 sec at 1000'C.
          4.2.3  Atomizing  time and temp:  10 sec at 2700*C.
          4.2.4  Purge gas:  Argon  (nitrogen should not be used).
          4.2.5  Wavelength:   357.9 nm.
          4.2.6  Background correction:  Not required.
          4.2.7  Other operating parameters should be  set as specified by the
                 particular instrument manufacturer.
     NOTE:   The above concentration values and instrument conditions are for a
             Perkin-Elmer  HGA-2100,  based  on  the   use  of  a 20-uL injection,
                                   7191 - 1
                                                          Revision      0
                                                          Date   September  1986

-------
            continuous-flow purge  gas,   and  nonpyrolytic  graphite.   Smaller
            sizes of  furnace  devices  or  those  employing  faster  rates  of
            atomization can be  operated  using lower atomlzatlon temperatures
            for shorter time periods than the above-recommended settings.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards:

          5.2.1  Stock solution:  Dissolve 1.923 g of chromium tr1oxide (CrOs,
     analytical reagent grade)  1n  Type  II  water,  acidify with redistilled
     HN03, and dilute to 1 liter.  Alternatively, procure a certified standard
     from a supplier and verify by comparison with a second standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used as
     calibration standards at the time of analysis.  These standards should be
     prepared to contain 0.5% (v/v) HN03; 1 ml of 30% ^2 and 1 ml of calcium
     nitrate solution, Section  5.2.3,  may  be  added to lessen interferences
     (see Section 3.0).

          5.2.3  Calcium  nitrate  solution:    Dissolve  11.8  g  of  calcium
     nitrate, Ca(N03)2*4H20  (analytical  reagent  grade),  1n Type II water and
     dilute to 1 liter.


 6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter  Three,  Section  3.1.3,  Sample  Handling and  Preservation.


 7.0  PROCEDURE

     7.1  Sample preparation;   The procedures   for  preparation  of the sample
 are  given in Chapter Three,  Section 3.2.

     7.2  See Method 7000,  Paragraph  7.3,   Furnace  Procedure.  The  calculation
 is given  in Method  7000,  Paragraph 7.4.


 8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method 7000.
 9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are  available  in Method 218.2  of  Methods
 for Chemical  Analysis of Water and Wastes.
                                   7191 - 2
                                                          Revision
                                                          Date   September 1986

-------
     9.2  The performance characteristics for an aqueous sample free of Inter-
ferences are:

     Optimum concentration range:  5-100 ug/L.
     Detection limit:  1 ug/L.

     9.3  The data shown In Table  1  were  obtained from records of state and
contractor laboratories.  The data are  intended  to show the precision of the
combined sample preparation and analysis method.


10.0  REFERENCES

1.   Methods for  Chemical  Analysis  of  Water  and Wastes,  EPA-600/4-82-055,
December 1982, Method 218.2.

2.   Gaskill, A., Compilation and Evaluation  of RCRA Method Performance Data,
Work  Assignment  No.  2,   EPA   Contract  No.  68-01-7075,   September  1986.
                                  7191 - 3
                                                         Revision
                                                         Date  September 1986

-------
                     TABLE 1.  METHOD PERFORMANCE DATA
Sampl e
Matrix
Paint primer
Contaminated soil
Oily lagoon soil
NBS SRM 1646 Estuarine sediment
EPA QC Sludge
NBS SRM 1085, Wear Metals in
lubricating oil
Preparation
Method
3050
3050
3050
3050
3050
3050
Laboratory
Replicates
2.7, 2.8 mg/g
12.0, 12.3 ug/g
69.6, 70.3 ug/g
42, 47 ug/ga
156 ug/gb
311, 356 ug/gc
aBias of -45 and -38% from expected, respectively.
bfiias of -24% from expected.
cB1as of +4 and +19% from expected, respectively.
                              7191 - 4
                                                     Revision      0
                                                     Date  September  1986

-------
                  METHOD 7191

CHROMIUM (ATOMIC  ABSORPTION. FURNACE TECHNIQUE)
5


il
Prepare
atandarda
                 7.1
                     I For
                     aampla
               preparation «e«
                   chapter 3.
                  aectlon 3.2
                7.Z  |

                Analyze ualng
                 Method 7000.
                 Section 7.3.
               calculation 7.4
              f     Stop      J
           7191  - 5
                                     Revision       0
                                     Date  September  1986

-------
                                 METHOD 7195

                   CHROMIUM,  HEXAVALENT (COPRECIPITATION)


1.0  SCOPE AND APPLICATION

     1.1  Method 7195 1s to  be  used  to  determine the concentration of dis-
solved hexavalent  chromium  [Cr(VI)]  In  Extraction  Procedure (EP) toxlclty
characteristic extracts and ground waters.  This method may also be applicable
to certain  domestic  and  Industrial  wastes,  provided  that  no Interfering
substances are present (see Paragraph 3.1 below).

     1.2  Method 7195 may be used to analyze samples containing more than 5 ug
of Cr(VI) per liter.   Either  flame or furnace atomic absorption spectroscopy
(Methods 7190 and 7191) can be used with copreclpltatlon.


2.0  SUMMARY OF METHOD

     2.1  Method 7195 is based on  the  separation  of Cr(VI) from solution by
coprecipltation of lead chromate  with  lead  sulfate  1n a solution of acetic
add.  After separation, the  supernate  [containing Cr(III)] is drawn off and
the precipitate is washed  to  remove  occluded  Cr(III).   The Cr(VI) is then
reduced and resolubillzed in nitric  acid  and quantified as Cr(III) by either
flame or furnace atomic absorption spectroscopy (Methods 7190 and 7191).


3.0  INTERFERENCES

     3.1  Extracts containing  either  sulfate  or  chloride 1n concentrations
above 1,000 mg/L should be diluted prior to analysis.


4.0  APPARATUS AND MATERIALS

     4.1  Filtering flask;  Heavy wall, 1-1 Her capacity.

     4.2  Centrifuge tubes;  Heavy  duty, conical, graduated, glass-stoppered,
10-mL capacity.

     4.3  Pasteur pipets;  Borosilicate glass, 6.8 cm.

     4.4  Centrifuge;   Any  centrifuge  capable  of  reaching  2,000  rpm and
accepting the centrifuge tubes described in Section 4.2 may be used.

     4.5  pH meter;  A wide  variety of instruments are commercially available
and  suitable for this work.

     4.6  Test tube mixer;  Any  mixer capable of imparting a thorough vortex
is acceptable.
                                   7195 -  1
                                                         Revision
                                                         Date  September 1986

-------
5.0  REAGENTS

     5.1  ASTM Type II water  (ASTM  D1193):    Water  should be monitored for
impurities.

     5.2  Lead nitrate solution;   Dissolve  33.1  g of lead nitrate,  Pb(N03)2
(analytical reagent grade), in Type II water and dilute to 100 mL.

     5.3  Ammonium sulfate solution;    Dissolve  2.7  g  of ammonium sulfate,
          (analytical reagent grade), 1n Type II water and dilute to 100 ml.
     5.4  Calcium nitrate  solution;    Dissolve  11.8  g  of calcium nitrate,
        *4H20 (analytical reagent grade), 1n Type II water and dilute to
100 ml (1 ml = 20 mg Ca) .

     5.5  Nitric acid;   Concentrated,  distilled reagent grade or spectrograde
quality.

     5.6  Acetic acid, glacial, 10%  (v/v):   Dilute 10 ml glacial acetic acid,
CH3COOH (ACS reagent grade), to 100 ml with Type II water.

     5.7  Ammonium hydroxide, 10% (v/v);    Dilute 10 ml concentrated ammonium
hydroxide, NH40H  (analytical reagent grade), to 100 ml with Type II water.

     5.8  Hydrogen peroxide, 30%;  ACS reagent grade.

     5.9  Potassium dichromate standard  solution;   Dissolve 28.285 g of dried
potassium dichromate,  I^C^Oy  (analytical reagent grade), in Type II water and
dilute to 1  liter (1 mL  = 10 mg Cr).

     5.10 Trivalent chromium working stock  solution;    To 50 ml of the potas-
sium dichromate  standard solution,  add  1 ml of 30% \\2®2 and  1 mL concentrated
HN03 and dilute  to  100 ml with  Type II water  (1 ml = 5.0 mg trivalent chro-
mium).   Prepare  fresh  monthly, or as needed.


6.0 SAMPLE  COLLECTION,  PRESERVATION, AND HANDLING

     6.1  All  samples  must   have  been   collected   using   a  sampling  plan  that
addresses the considerations  discussed  in Chapter Nine  of  this manual.

     6.2  Since  the stability of   Cr(VI)  in   EP   extracts   is  not  completely
 understood  at this  time,  the  analysis should  be   carried out   as  soon  as
possible.

     6.3   To retard the chemical  activity  of  hexavalent  chromium,  samples and
 extracts  should  be stored at   4*C   until analyzed.    The  maximum  holding  time
 prior  to analysis 1s 24 hr.
                                   7195 - 2
                                                          Revision
                                                          Date  September 1986

-------
7.0  PROCEDURE

     7.1  Transfer a 50-mL portion of  the  sample  to a 100-mL Griffin beaker
and adjust to a pH of 3.5 + 0.3 by adding volumes of 10% acetic acid dropwlse.
Proceed Immediately to Step 7.2,  taking  no  longer than 15 mln between these
steps.
     NOTE:  Care must be exercised not to take  the  pH below 3.  If the pH is
            inadvertently lowered to <3, 10%  NfyOH should be used to readjust
            the pH to 3.5 + 0.3.

     7.2  Pi pet a 10-mL aliquot of the adjusted sample into a centrifuge tube.
Add 100 uL of the lead nitrate solution, stopper the tube, mix the sample, and
allow to stand for 3 min.

     7.3  After the formation of lead chromate, to help retain Cr(III) complex
in solution, add 0.5 ml glacial acetic acid, stopper, and mix.

     7.4  To provide adequate lead sulfate  for coprecipitation, add 100 uL of
ammonium sulfate solution, stopper, and mix.

     7.5  Place the stoppered centrifuge  tube  in the centrifuge, making sure
that the tube is properly  counterbalanced.    Start the centrifuge and slowly
increase the speed to 2,000 rpm  in  small  increments over a period of 5 min.
Hold at 2,000 rpm for 1 min.
     NOTE:  The speed of the  centrifuge  must  be  increased slowly to ensure
            complete coprecipitation.

     7.6  After centrifuging, remove  the  tube  and  withdraw and discard the
supernate   using  either  the  apparatus  detailed  in  Figure  1  or  careful
decantation.  If using the vacuum apparatus, the pasteur pipet is lowered into
the tube and the supernate  1s  sucked  over   into  the filtering flask.  With
care, the supernate can be withdrawn  to within approximately 0.1 ml above the
precipitate.  Wash the precipitate with  5  ml  Type  II water and repeat steps
7.5 and 7.6; then proceed to 7.7.

      7.7  To the remaining precipitate, add  0.5  ml  concentrated HN03, 100 uL
30% H202, and 100 uL  calcium   nitrate  solution.     Stopper  the tube  and mix,
using a vortex  mixer  to  disrupt  the  precipitate  and  solubllize  the lead
chromate.   Dilute to  10  ml,   mix,  and  analyze   in   the  same manner as the
calibration standard.

      7.8  Flame atomic absorption;  At  the  time of  analysis,  prepare a  blank
and a series of at  leastfourcalibration standards from the  Cr(III)  working
stock that  will adequately bracket the  sample and  cover  a concentration  range
of 1  to 10  mg Cr/L.   Add  to   the blank  and each standard, before  diluting to
final volume,  1 ml  30% \\2®2, 5 ml concentrated HNOs, and 1 ml  calcium nitrate
solution  for each  100 ml   of   prepared   solution.   These  calibration  standards
should  be prepared  fresh weekly,  or  as  needed.  Refer  to Method  7090 for more
detail.
                                   7195 - 3
                                                          Revision      0
                                                          Date  September  1986

-------
     7.9  Furnace atomic absorption;  At the time of analysis,  prepare a blank
and a series of at leastfour  calibration standards from the  Cr(III) working
stock that will adequately bracket the  sample and cover a concentration range
of 5 to 100 ug Cr/L.  Add  to  the blank and each standard, before diluting to
final volume, 1 ml 30% ^03, 5  ml concentrated HN03, and 1 ml  calcium nitrate
solution for each 100 ml  of  prepared  solution.  These calibration standards
should be prepared fresh weekly, or as  needed.  Refer to Method 7191 for more
detail.

     7.10 Verification;

          7.10.1  For every sample  matrix  analyzed, verification is required
     to ensure that neither a  reducing condition nor chemical  interference is
     affecting precipitation.  This must be accomplished by analyzing a second
     10-mL aliquot of  the  pH-adjusted  filtrate  that  has  been spiked with
     Cr(VI).  The amount of spike  added should double the concentration found
     in the original aliquot.   Under  no  circumstance should the increase be
     less than 30 ug/L Cr(VI).  To  verify the absence of an interference, the
     spike recovery must be between 85% and 115%.

          7.10.2  If addition of  the  spike  extends the concentration beyond
     the calibration curve, the analysis solution should be diluted with blank
     solution and the calculated results adjusted accordingly.

          7.10.3  If  the  result  of  verification  indicates  a  suppressive
     interference, the sample should be diluted and  reanalyzed.  If necessary,
     use furnace atomic absorption to achieve the optimal concentration range.

          7.10.4  If  the  interference  persists  after  sample  dilution, an
     alternative method  (Method  7197,  Chelation/Extraction,  or Method 7196,
     Colorimetric) should be used.

     7.11  Acidic extracts that yield  recoveries  of  less than 85%  should be
retested to  determine if the  low   spike   recovery   is  due to the presence of
residual reducing agent.    This  determination  shall  be  performed by first
making  an aliquot  of  the  extract  alkaline   (pH   8.0-8.5)  using  1 N sodium
hydroxide and  then  respiking and analyzing.   If  a spike recovery of  85-115% is
obtained  in  the  alkaline aliquot of an acidic  extract that  initially  was found
to contain  less  than  5  mg/L   Cr(VI),  one   can  conclude  that the  analytical
method  has  been  verified.


8.0 QUALITY CONTROL

     8.1   All  quality  control data  should  be  maintained  and available for  easy
reference or inspection.

     8.2   Calibration  curves  must  be   composed  of   a  minimum of a blank and
three  standards.   A  calibration   curve  should be made  for  every hour of
continuous sample analysis.
                                   7195 - 4
                                                          Revision      0
                                                          Date  September 1986

-------
            8.3  Dilute samples  if  they  are  more  concentrated  than  the highest
       standard or  if they fall on the plateau of a calibration curve.

            8.4  Employ a minimum  of  one  blank  per  sample  batch to determine if
       contamination or any memory effects are occurring.

            8.5  Verify calibration  with  an  independently  prepared check standard
       every  15 samples.

            8.6  Run one spike duplicate sample  for  every  10 samples.  A duplicate
       sample 1s a  sample brought  through the whole sample preparation and analytical
       process.

            8.7  The method of   standard  additions   (see  Method   7000, Section  8.7)
       shall  be used for the  analysis  of  all EP extracts, on all  analyses submitted
       as part of  a del 1 sting petition,  and  whenever  a new sample matrix is being
       analyzed.


       9.0 METHOD PERFORMANCE

            9.1   Precision  and accuracy data  are available 1n Method 218.5 of Methods
       for Chemical Analysis  of  Water  and Wastes.


       10.0  REFERENCES

       1.  Methods for  Chemical   Analysis   of  Water   and Wastes, EPA-600/4-82-055,
       December  1982,  Method  218.5.

•
                                          7195 - 5
                                                                 Revision
                                                                 Date  September 1986

-------
                                            METHOD 7195

                            HEXAVALENT CHROMIUM:  COPRECIPITATION METHOD
    7. 1
   Transfer sample
      to beaker;
      adjust PH
 7.2
                             7.6
                                   Remove
                              tube;  discard
                            supernate:  wash
                              precipitate;
                            repeat 7.5.  7.6
Pipet adjusted cample
Into centrifuge tube;
   add l«ad nitrate
 solution:  mix :  let
        stand
    7.3
                                                        7.7
                                   Add cone
                               	1   HNOi.
                               SOX  HjOz and
                           •calcium nitrate
                             solution:  mix;
                            dilute:  analyze
     Add glacial
     acetic acid;
         mix
    7.4
                           7.9
 Prepare blank and
     series of
 standards covering
concentration range
  or S to 100 ug
     Cr/lUer
    Add ammonium
      sulfate
   solution:  mix
                              7.9
                                                 Furnace
                                                                      Flame
                                                        Which type of
                                                      atomic absorption
                                                          is used?
                                                                               7.8
 Prepare blank and
      series of
 standards covering
concentration range
   of 1 to 10 mg
     Cr/liter
                                    Add 30X
   cone HNOj.  and
  calcium nitrate
     solution to
    each;  analyze
                                                                                  7.e
         Add 30X
          H,0t.
    cone HNOj  and
  calcium nitrate
     solution to
    each:  analyze
    7.3
    Place tube in
     centrifuge:
     centrifuge
                             7.10.1
                                     Verify
                               by analyzing
                             second aliquot
                                 of spiked
                                  filtrate
     o
                                         7195 - 6
                                                                    Revision       0
                                                                    Date   September 1986

-------
                                       METHOD 7195

                        HEXAVALENT CHROMIUM: COPHECIPITATION METHOD
                                        (Continued)
             ^^i w
Is suppresslve >v
 Interference
  Indicated?
                          Dilute sample
                          and reanalyze
 7.12 |

       If no valid
results,  and chromium
  more than threshold
 amount of hexavalent
    chromium,  sample
exihiblte EP toxiclty
                                                                                  Analytic
                                                                                   method
                                                                            verified: waste
                                                                              not  hazardous
                                                 f     Stop      J
                                    7195 - 7
                                                               Revision       Q
                                                               Date   September 1986

-------
                                 METHOD 7196A

                      CHROMIUM, HEXAVALENT  (COLORIMETRIC)
1.0  SCOPE AND APPLICATION

      1.1   Method 7196  is used to  determine  the concentration  of  dissolved
hexavalent  chromium  [Cr(VI)]  in  EP/TCLP characteristic  extracts and  ground
waters.  This method may also be applicable to certain domestic and industrial
wastes, provided that no interfering substances are present (see Paragraph 3.1
below).

      1.2   Method 7196 may be used to analyze samples containing from 0.5 to
50 mg of Cr(VI) per liter.


2.0  SUMMARY OF METHOD

      2.1   Dissolved hexavalent chromium, in the absence of interfering amounts
of  substances  such as  molybdenum,  vanadium,  and  mercury,  may  be determined
colorimetrically by reaction with diphenylcarbazide  in  acid  solution.   A red-
violet color of unknown composition is produced.  The reaction  is very sensitive,
the absorbancy  index  per gram  atom of chromium being about  40,000 at  540 nm.
Addition of an excess of diphenylcarbazide yields the red-violet product, and its
absorbance is measured photometrically at 540 nm.


3.0  INTERFERENCES

      3.1   The chromium  reaction with diphenylcarbazide  is  usually  free from
interferences.   However,  certain  substances  may interfere  if  the  chromium
concentration is relatively low.  Hexavalent molybdenum and mercury salts also
react  to form  color  with  the reagent;  however,  the  red-violet intensities
produced  are  much  lower  than those   for  chromium   at  the  specified  pH.
Concentrations  of  up  to 200 mg/L of molybdenum  and  mercury  can  be tolerated.
Vanadium interferes strongly, but concentrations up to 10 times that of chromium
will not cause trouble.

      3.2   Iron  in  concentrations  greater than  1  mg/L may  produce  a yellow
color, but the  ferric  iron color is  not  strong and difficulty is not normally
encountered if  the absorbance  is measured photometrically at the appropriate
wavelength.


4.0  APPARATUS AND MATERIALS

      4.1   Colorimetric equipment:  One of the following is  required:   Either
a spectrophotometer,  for use at 540 nm, providing  a light path  of 1  cm or longer,
or  a filter photometer,  providing a  light path of 1  cm or longer and equipped
with  a greenish-yellow filter  having  maximum  transmittance  near
540 nm.
                                   7196A -  1                       Revision 1
                                                                  July 1992

-------
5.0  REAGENTS

      5.1   Reagent water:  Reagent water  should  be monitored  for
impurities.

      5.2   Potassium dichromate  stock  solution:   Dissolve 141.4  mg  of dried
potassium dichromate, K2Cr,07  (analytical  reagent  grade),  in reagent water and
dilute to 1 liter (1 ml = 50 ug Cr).

      5.3   Potassium dichromate standard solution:  Dilute 10.00 ml potassium
dichromate stock solution to 100 ml (1 ml = 5 ug Cr).

      5.4   Sulfuric acid, 10% (v/v):  Dilute 10 ml of distilled reagent grade
or spectrograde quality sulfuric acid, H2S04,  to  100  ml with reagent water.

      5.5   Diphenylcarbazide solution:  Dissolve 250 mg 1,5-diphenylcarbazide
in 50 ml acetone.  Store  in a brown bottle.  Discard when the solution becomes
discolored.

      5.6   Acetone (analytical reagent grade): Avoid or redistill material that
comes in containers with  metal or metal-lined caps.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must  have been  collected using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   Since  the  stability  of  Cr(VI)  in  extracts  is  not  completely
understood at this time,  the analysis  should be carried out as soon  as possible.

      6.3   To retard the chemical activity of hexavalent chromium, the samples
and extracts should be stored at  4°C until analyzed.  The maximum holding time
prior to analysis of the  samples or extracts  is 24 hr.  The 24 hr holding time
begins after extraction.


7.0  PROCEDURE

     7.1    Color development and measurement: Transfer 95 mL of the extract to
be tested to a 100-mL volumetric  flask.  Add  2.0 mL diphenylcarbazide solution
and mix.   Add H2S04  solution  to give a  pH  of 2 + 0.5, dilute  to  100 mL with
reagent water, and let stand 5 to  10 min for  full color development.  Transfer
an appropriate portion of the  solution to a 1-cm absorption cell  and measure its
absorbance at 540 nm.   Use reagent  water as a reference.  Correct the absorbance
reading of the sample by  subtracting  the absorbance of a blank carried through
the method (see Note below).  An  aliquot of the sample containing  all reagents
except diphenylcarbazide  should  be prepared and used to correct the sample for
turbidity  (i.e., a turbidity blank).  From the corrected absorbance, determine
the mg/L of chromium present by  reference to  the calibration curve.

      NOTE: If  the  solution  is  turbid  after  dilution to  100 mL  in Step 7.1,
            above,  take  an  absorbance  reading  before  adding  the carbazide

                                   7196A - 2                       Revision 1
                                                                   July  1992

-------
            reagent and  correct  the absorbance  reading  of the  final  colored
            solution by subtracting the absorbance measured previously.

      7.2   Preparation of calibration  curve:

            7.2.1  To compensate for possible  slight losses of chromium during
      digestion  or other  operations  of  the  analysis,   treat  the  chromium
      standards  by  the same procedure  as  the  sample.   Accordingly, pipet  a
      chromium standard  solution  in measured  volumes  into 250-mL  beakers  or
      conical flasks  to generate standard  concentrations  ranging from 0.5 to
      5 mg/L Cr(VI) when diluted to the appropriate volume.

            7.2.2   Develop the  color  of  the standards  as  for  the  samples.
      Transfer a suitable portion  of each colored  solution  to a 1-cm absorption
      cell and measure the absorbance  at  540  nm.  As reference,  use  reagent
      water.  Correct  the  absorbance readings  of the  standards by subtracting
      the absorbance of a reagent blank carried through the method.  Construct
      a calibration curve by plotting corrected absorbance values against mg/L
      of Cr(VI).

     7.3    Verification:

            7.3.1  For  every sample matrix analyzed,  verification  is required to
      ensure that  neither  a reducing  condition nor chemical  interference  is
      affecting  color  development.   This must be accomplished by analyzing  a
      second 10-mL aliquot  of the pH-adjusted filtrate that has  been spiked with
      Cr(VI).  The amount of spike added should double the concentration found
      in the original  aliquot.  Under  no circumstances should  the increase be
      less than 30 /ig  Cr(VI)/liter.  To verify the absence of an interference,
      the spike recovery must be between 85% and 115%.

            7.3.2  If addition of the spike  extends the  concentration beyond the
      calibration  curve,  the  analysis  solution  should be diluted  with blank
      solution and the calculated results adjusted accordingly.

            7.3.3   If the  result of  verification  indicates  a  suppressive
      interference, the sample should be diluted and reanalyzed.

            7.3.4   If  the interference persists  after  sample   dilution,  an
      alternative  method  (Method  7195,   Coprecipitation,  or  Method  7197,
      Chelation/Extraction) should be used.

      7.4   Acidic extracts that  yield recoveries of  less than  85% should  be
retested  to  determine  if  the  low  spike recovery is  due  to the  presence  of
residual reducing agent.  This determination shall be performed by first making
an aliquot of the extract alkaline  (pH  8.0-8.5) using 1 N sodium hydroxide and
then respiking and analyzing.  If a spike recovery of 85-115%  is obtained in the
alkaline aliquot of an acidic extract that  initially was found to contain less
than  5  mg/L Cr(VI),  one  can  conclude that  the  analytical  method has  been
verified.
                                   7196A  - 3                       Revision 1
                                                                  July 1992

-------
      7.5   Analyze all extracts, all  samples  analyzed  as  part  of a deli sting
petition, and all  samples that suffer from matrix interferences by the method of
standard additions (see Method 7000, Section 8.7).


8.0  QUALITY CONTROL

      8.1   All quality control data should be maintained and available for easy
reference or inspection.  Refer to Chapter One for more information.

      8.2   Dilute  samples  if  they are  more concentrated  than   the  highest
standard or if they fall on the plateau of a calibration curve.

      8.3   Employ  a  minimum of  one  blank per  sample  batch to  determine if
contamination or any memory effects are occurring.

      8.4   Verify calibration  with an independently prepared  check standard
every 15 samples.

      8.5   Run one matrix  spike replicate or one replicate  sample  for every ten
samples.   A duplicate  sample is  a sample brought  through  the  whole  sample
preparation and analytical  process.  Refer to Chapter One for more information
concerning matrix spikes and matrix spike duplicates.

      8.6   The method of standard additions (see Method 7000,  Section 8.7) shall
be used for the analysis of all extracts, on all analyses submitted as part of
a delisting petition, and whenever a new sample matrix is being analyzed.


9.0  METHOD PERFORMANCE

      9.1   The data  shown  in  Table 1  were obtained  from records  of state and
contractor  laboratories.   The data are intended  to  show the  precision  of the
combined sample preparation and analysis method.


10.0  REFERENCES

1.    Methods  for Chemical  Analysis  of  Water and  Wastes,  EPA-600/4-82-055,
December 1982, Methods 218.4 and  218.5.

2.    Gaskill, A., Compilation and  Evaluation of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075,  September 1986.
                                   7196A - 4                       Revision  1
                                                                   July  1992

-------
                      TABLE 1. METHOD PERFORMANCE DATA
Sample
Matrix
Preparation
  Method
Laboratory
Replicates
Wastewater treatment
  sludge                    Not known

Sediment from chemical
 storage area                 3060
                    0.096,  0.107 ug/g
                        115,  117 ug/g
                                  7196A  -  5
                                     Revision 1
                                     July 1992

-------
                                                    METHOD  7196A
                                    CHROMIUM,  HEXAVALENT  (COLORIMETRIC)
       Start
   ? 1 Transfer
    extract to
     flask,add
 diphenyIcarbazide
 solution,and nix
     for  color
    development
   7 1  Add H,SO«
solution.dilute,let
 stand,measure  the
correct absorbance
    reading,and
   determine Cr
     present
  7 2.1  Treat Cr
 standards by the
 same procedure as
  sample,pipet Cr
 standard solution
    into beaker
722  Develop color
  for  standards,
measure  and correct
 reading,construct
 calibration curve
  731 Analy2e a
 second aliquot of
    pK adjusted
  filtrate spiked
  with Cr(VI) for
   verification
7.3.2 Dilute
•piked aanplt
with blink
solution,
adjutt recult*

Ye«


   Do**
• pike cone
xeeed calibr
  curve?
     I.
 •upre»»iv»
interferene
  indicated
7.3.3  Dilute
 •••pie end
                                          7.4  Prepare an
                                            alkaline
                                          aliquot «itk IN
                                            NaOH,«pike
                                          •a>ple,analyie
                         7  4  Analytical
                       method i» verified
                              Stop
                                                      7196A -  6
                                                                     Revision 1
                                                                     July 1992

-------
                                 METHOD 7197

                 CHROMIUM, HEXAVALENT (CHELATION/EXTRACTION)


1.0  SCOPE AND APPLICATION

     1.1  Method  7197  1s  approved  for  determining  the  concentration  of
dissolved hexavalent chromium [Cr(VI)]  1n  Extraction Procedure (EP) toxicity
characteristic extracts and ground waters.  This method may also be applicable
to certain  domestic  and  Industrial  wastes,  provided  that  no interfering
substances are present (see Paragraph 3.1).

     1.2  Method 7197 may be used  to analyze  samples containing  from 1.0 to
25 ug of Cr(VI) per liter.


2.0  SUMMARY OF METHOD

     2.1  Method 7197 1s  based  on  the  chelation of hexavalent chromium with
ammonium  pyrrolidlne  dlthlocarbamate   (APDC)   and  extraction  with  methyl
Isobutyl ketone (MIBK).   The extract  is  aspirated into the flame of an atomic
absorption spectrophotometer.


3.0  INTERFERENCES

     3.1  High concentrations of other metals may Interfere.


4.0  APPARATUS AND MATERIALS

     4.1  Atomic  absorption  spectrophotometer;    Single  or  dual  channel,
single-  ordouble-beaminstrument,having   a   grating  monochromator,
photomultiplier detector,  adjustable slits,  and  provisions  for background
correction.

     4.2  Chromium hollow cathode  lamp.

     4.3  Strip-chart recorder  (optional).


5.0  REAGENTS

     5.1  ASTM Type II water  (ASTM   D1193):    Water  should be monitored for
Impurities.

     5.2  Ammonium  pyrrolidine  dithlocarbamate  (APDC)  solution:   Dissolve
1.0  g APDC in Type II water and dilute to  100 mL.  Prepare fresh daily.

     5.3  Bromphenol blue Indicator  solution;   Dissolve 0.1 g bromphenol blue
1n 100 mL 50% ethanol.
                                   7197 -  1
                                                         Revision
                                                         Date  September 1986

-------
     5.4  Potassium dlchromate standard  solution  I  (1.0  ml  =   100 ug  Cr):
Dissolve 0.2829 g pure dried  potassium  dichromate, I^C^O/,  in Type II water
and dilute to 1,000 mL.

     5.5  Potassium dichromate standard solution  II  (1.0  mL  =  10.0 ug  Cr):
Dilute 100 mL chromium standard solution I to 1 liter with Type II water.

     5.6  Potassium dichromate standard solution  III  (1.0  mL =  0.10 ug  Cr):
Dilute 10.0 ml chromium standard solution II to 1 liter with Type  II water.

     5.7  Methyl isobutyl ketone (MIBK),  analytical  reagent grade:  Avoid  or
redistill material that comes in contact with metal or metal-lined caps.

     5.8  Sodium hydroxide solution, 1 M:   Dissolve to 40 g sodium hydroxide,
NaOH (ASC reagent grade), in Type II water and dilute to 1 liter.

     5.9  Sulfuric acid, 0.12 M:  Slowly add 6.5 ml distilled reagent grade  or
spectrograde-quality sulfuric acid, ^$04, to  Type  II  water and dilute to 1
liter.
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  All  samples must  have  been  collected  using  a sampling plan that
addresses the  considerations discussed in Chapter Nine of this manual.

     6.2  Because  the stability of   Cr(VI)   in  EP  extracts is not completely
understood at  this time,  the chelation and extraction should be carried out as
soon as possible.

     6.3  To  retard the chemical  activity  of hexavalent chromium, the samples
and extracts  should be stored  at  4*C until analyzed.


7.0  PROCEDURE

     7.1   Pi pet  a volume  of  extract containing  less  than   2.5  ug  chromium
 (100 mL maximum)  Into  a   200-mL   volumetric flask   and  adjust the  volume to
approximately 100 mL.

     7.2   Prepare a blank and   sufficient  standards  and adjust the  volume of
each to approximately  100 mL.

     7.3   Add 2 drops  of  bromphenol   blue  indicator solution.   (The adjustment
of pH  to  2.4,  Step  7.4,   may   be  made  with   a  pH  meter  instead of using an
 indicator.)

     7.4   Adjust the pH  by addition of 1 M NaOH solution dropwise  until  a blue
 color  persists.  Add 0.12 M  H?S04  dropwise until the blue color just  disap-
pears  in  both the standards and sample.    Then  add  2.0 mL of 0.12 M H2S04 in
excess.   The  pH at this  point  should be  2.4.


                                   7197 - 2
                                                         Revision      0
                                                          Date  September 1986

-------
     7.5  Add 5.0 ml APDC solution and  mix.    The pH should then be approxi-
mately 2.8.

     7.6  Add 10.0 ml MIBK and shake vigorously for 3 mln.

     7.7  Allow the layers to separate and  add Type II water until the ketone
layer 1s completely 1n the neck of the flask.

     7.8  Aspirate the ketone  layer  and  record  the  scale reading for each
sample and standard against  the  blank.    Repeat,  and average the duplicate
results.

     7.9  Determine the mg/Hter of Cr(VI) in each sample from a plot of scale
readings of standards.  A  working  curve  must  be  prepared with each set of
samples.

    7.10  Verification;

          7.10.1  For every sample  matrix  analyzed, verification 1s required
     to ensure that neither a  reducing condition nor chemical interference is
     affecting chelatlon.  This must be accomplished by analyzing a second 10-
     mL aliquot of the pH-adjusted filtrate  that has been spiked with Cr(VI).
     The amount of spike added  should  double  the concentration found 1n the
     original aliquot.  Under  no  circumstances  should  the Increase be less
     than 30 ug/L Cr(VI).  To verify the  absence of an interference, the spike
     recovery must be between 85% and 115%.

          7.10.2  If addition of  the   spike  extends the concentration beyond
     the calibration curve, the analysis  solution should be diluted with blank
     solution and the calculated  results  adjusted accordingly.

           7.10.3  If  the  result  of   verification   Indicates  a  suppressive
     interference, the  sample should be diluted and  reanalyzed.

           7.10.4  If  the  Interference  persists  after  sample   dilution, an
     alternative  method   (Method  7195,    Coprecipitation,   or   Method  7196,
     Color1metric) should  be used.


      7.11  Acidic extracts that yield   recoveries  of  less  than  85%  should be
 retested to  determine  1f the   low  spike   recovery   Is  due  to  the presence of
 residual  reducing agent.    This  determination   shall  be   performed by first
 making an  aliquot  of  the extract   alkaline   (pH   8.0-8.5)  using  1 N  sodium
 hydroxide  and  then resplklng and  analyzing.   If  a  spike recovery  of  85-115% is
 obtained in  the alkaline aliquot  of an  acidic  extract that  Initially  was found
 to contain less than   5 mg/L  Cr(VI),  one  can   conclude  that the  analytical
 method has been verified.
                                   7197 - 3
                                                          Revision       0
                                                          Date   September  1986

-------
8.0  QUALITY CONTROL

     8.1  All quality control data should be maintained and available for easy
reference or Inspection.

     8.2  Calibration curves must be  composed  of  a  minimum  of a blank and
three standards.   A  calibration  curve  should  be  made  for  every hour of
continuous sample analysis.

     8.3  Dilute samples  if  they  are  more  concentrated  than  the highest
standard or 1f they fall on the plateau of a calibration curve.

     8.4  Employ a minimum  of  one  blank  per  sample  batch to determine 1f
contamination or any memory effects are occurring.

     8.5  Verify calibration  with  an  Independently  prepared check standard
every 15 samples.

     8.6  Run one spike duplicate sample  for  every  10 samples.  A duplicate
sample 1s a sample brought through the whole sample preparation and analytical
process.

     8.7  The method of  standard  additions  (see  Method  7000, Section 8.7)
shall be used for the analysis  of  all EP extracts, on all analyses submitted
as part of a del 1 sting  petition,  and  whenever  a new sample matrix 1s being
analyzed.


9.0  METHOD PERFORMANCE

     9.1  Precision and accuracy data are available 1n Method 218.4 of Methods
for  Chemical Analysis of Water and Wastes.


10.0 REFERENCES

1.   Methods for   Chemical   Analysis  of  Water   and Wastes,  EPA-600/4-82-055,
December  1982,  Method 218.4.
                                   7197 - 4
                                                          Revision
                                                          Date  September 1986

-------
                             METHOD 7197

             HEXAVALENT CHROMIUM (CHELATION/EXTHACTION)
(      Start      1
  7. 1
  Plpet extract
   Into flask:
  adjust volume
  7.2
        Prepare
       blank and
      standards:
   adjust volume
       of each
  7.3
  Add bromphenol
  blue Indicator
     solution
  7 .A
Allow layers to
 separate:  add
 Type II water
  Adjust pH by
  adding NaOH;
   add
                                                      7.8
       Aspirate
       ketone
  layer;  record
scale readings:
repeat:  average
     results
                            7197 - 5
                                                      Revision       0
                                                      Date  September 1986

-------
                                         METHOD 7197

                         HEXAVALENT CHROMIUM  (CHEUATION /EXTRACTION)
                                         (Continued)
JLU
      Determine
 Cr (IV)  in each
sample:  prepare
 working curves
 7.10.1  Verify
       I every
  sample matrix
  by analyzing
 second aliquot
spiked filtrate
    Is cone.
     beyond
  calibration
     curve?
       analysis
  solution with
blank solution;
 adjust results
   Is there a
  suppresslve
 interference?
                           Dilute  sample
                           and  reanalyze
                                                   Use alternative
                                                       method
   Analytical
method verified
 -waste Is not
   hazardous
                              If no valid
                        results and chromium
                         concentration over
                          threshold limits.
                          sample exhibits
                            EP toxlclty
                                                  f     Stop      J
                                      7197 - 6
                                                                Revision       0
                                                                Date  September  1986

-------
                                 METHOD 7198

           CHROMIUM, HEXAVALENT (DIFFERENTIAL PULSE POLAROGRAPHY)


1.0  SCOPE AND APPLICATION

     1.1  This method is  used  to  determine  the concentration of hexavalent
chromium [Cr(VI)] in natural and waste waters and in EP extracts.

     1.2  The method  can  quantitate  chromium  1n  concentrations  of  up to
1.0 mg/L to 5.0 mg/L, depending  on  the mercury drop size.  Higher concentra-
tions can be determined by dilution.

     1.3  The  lower  limit  of  detection  for  Cr(VI)  1s  10  ug/L  for the
instrumental conditions given in this method.  The limit of detection could be
easily lowered by changing these conditions.


2.0  SUMMARY OF METHOD

     2.1  Method 7198 measures the peak current produced from the reduction of
Cr(VI) to Cr(III) at a dropping  mercury electrode during a differential pulse
voltage ramp.

     2.2  The method described herein uses 0.125  M NH40H-0.125 M NH4C1 as the
supporting electrolyte.  In this electrolyte, Cr(VI) reduction results in peak
current occurring at the peak potential (Ep) of -0.250 V vs. Ag/AgCl.

     2.3  Alternative supporting electrolytes, such as those given in Table 1,
may be used.

     2.4  The technique of standard additions  must  be used to quantitate the
Cr(VI) content.


3.0  INTERFERENCES

     3.1  Copper ion at concentrations higher than the Cr(VI) concentration is
a potential Interference due to peak overlap when using the 0.125 M ammoniacal
electrolyte.  Increasing  the  ammoniacal  electrolyte  concentration to 0.5 M
shifts  the  copper peak  cathodically    (Ep   =  -0.4  V),  eliminating  the
Interference at a copper-to-chromium ratio of 10:1 (Figure 1).

     3.2  Reductants such as ferrous  iron,  sulflte,  and sulflde will reduce
Cr(VI) to Cr(III);  thus it  is  imperative  to  analyze the samples as soon as
possible.

4.0  APPARATUS AND  MATERIALS

     4.1  Polarographic 1nstrumentation;   Capable  of performing differential
pulse analyses,  including recorder or plotter.


                                  7198 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
  2.0-
  1.5-
                  Current x 102 nA
                  18 Jan 82 No. 1
                  Sample: DPP
                  Initial E: -0.100 V
                  Final E: -0.450 V

                  Peak 1: -0.292 V
                         2.047 E2nA
 4.0-
  3.0-
Current x 102 nA
18 Jan 82  No. 2
Sample: DPP
Initial E: -0.100V
Final E: -0.450 V

Peak 1: -0.256 V
       2.680 E1 nA

Peak 2: -0.396 V
       9.740E1 nA
       -0.2    -0.4
      -0.2   -0.4
 A. 20 ppm Cu, 2.5 ppm Cr, 0.1 N buffer.
B. 20 ppm Cu, 2.5 ppm Cr, 0.5 N buffer.
     Figure 1. Two polarograms illustrating shift in copper peak at higher ammoniacal
electrolyte concentrations.
                              7198 - 2
                                                          Revision       0
                                                          Date   September  1986

-------
                TABLE 1.   POLAROGRAPHY OF  HEXAVALENT CHROMIUM
      Supporting electrolyte
Peak potential  (vs. SCE)
1 M NaOH
1 M Pyr1d1ne, 1 M NaOH
1 M NH4OH, 1 M NH4C1
0.1 M NH40H, 0.1 M (^4)2 Tartrate
0.2 M KC1, 0.3 M Tr1ethanolam1ne, pH 9
1 M Na2S04
0.1 M NH4OH, 0.1 M NH4C1
        -0.85
        -1.48
        -0.36
        -0.244
        -0.28
        -0.23
        -0.25
                                  7198 - 3
                                                         Revision      0
                                                         Date  September 1986

-------
     4.2  Dropping  mercury  electrode   assembly:      Capable  of  performing
differential pulse analyses.

     4.3  Counter electrode:  Platinum wire.

     4.4  Reference electrode;  Ag/AgCl  or  SCE,   with a slow-leakage fritted
tip (unfired Vycor).

     4.5  Nitrogen gas and cell outgassing assembly.

     4.6  Micropipets and disposable tips.


5.0  REAGENTS

     5.1  ASTM Type II water   (ASTM  D1193):     Water  should be monitored for
impurities.

     5.2  Chromium standard solution I, 1.0 ml  =  100  ug Cr:  Should be made
daily from a 1,000-ppm standard stock solution made with Type II water.

     5.3  Chromium standard solution II,  1.0 ml  =  10  ug Cr:  Should be made
daily from a 1,000-ppm standard stock solution made with Type II water.

     5.4  Chromium standard solution III,  1.0  ml  =  1  ug  Cr:  Dilute 10 ml
chromium standard solution  II  to 100 ml with Type II water.

     5.5  Ammoniacal  electrolyte, 2.5 N:   Dissolve  33.3  g of NfyCl  in 150 ml
of  Type II  water, add 42.2  ml  of concentrated NfyOH, and dilute to 250 ml.

     5.6  Concentrated nitric  acid;    Acid  should  be  analyzed to  determine
levels of  impurities.ITimpurities  are  detected,  all analyses  should be
blank-corrected.
 6.0  SAMPLE COLLECTION,  PRESERVATION,  AND HANDLING

      6.1   All  samples must  have  been  collected   using   a  sampling plan that
 addresses the  considerations discussed in Chapter Nine  of  this manual.

      6.2   Stability of  Cr(VI)   is  not  completely  understood   at this time.
 Therefore, the analysis  should  be carried out  as soon as possible.

      6.3   If the analysis cannot be performed  within  24 hr,  take  an aliquot  of
 the sample and add  a known amount of  Cr(VI)   (0.1 mg/L   for natural   waters,
 1 mg/L for wastewaters,   and  5  mg/L  for  EP extracts).  Analyze this known
 additional sample at the same time the sample  is analyzed  to determine  whether
 Cr(VI) was reduced during storage.

      6.4   To retard the   chemical  activity of Cr(VI),   the sample should  be
 transported and stored at 4*C until time of analysis.
                                   7198 - 4
                                                          Revision
                                                          Date  September 1986

-------
7.0  PROCEDURE

     7.1  Soak the voltammetric cells overnight  in  1  +  1 HNC-3 and/or 1 + 1
aqua regia.

     7.2  Rinse the electrode assembly with Type II water,  then with 1 N HMOs,
and finally with Type II water prior to and in between sample analyses.

     7.3  The  instrument  should  be  set  using  the  following instrumental
parameters.

          7.3.1  Mode:  Differential pulse.
          7.3.2  Scan rate:  2 mV/sec.
          7.3.3  Drop time:  1 sec.
          7.3.4  Initial potential:  -0.05 V + 0.05 V vs. Ag/AgCl.
          7.3.5  Final potential:  -0.50 V + 0.10 V vs. Ag/AgCl.
          7.3.6  Pulse height:  0.05 V.
          7.3.7  Deaeratlon time:  240 sec  or  less initially, 30 sec between
                 standard additions.

     7.4  Analysis:

          7.4.1  Pipet a volume of  sample  containing  less than 10 ug Cr(VI)
     into a voltammetric cell  (the  maximum volume depends on the voltammetric
     cell volume, usually 10 ml).

          7.4.2  Add 0.5 ml of the ammoniacal electrolyte and adjust volume to
     10 ml with Type II water.

          7.4.3  Place the electrode assembly in  the  solution and outgas with
     nitrogen  for at least 120 sec.

          7.4.4  Engage the electrode  assembly  to the polarographic  analyzer
     and displace at least 10 mercury  drops before initiating the voltage ramp
     and obtaining  the polarogram.

           7.4.5  Figure 2 gives  typical differential  pulse  polarograms.

     7.5   Prior to  the  analysis  of   any   samples,   and  during analysis at a
 frequency  of at least once every  10   samples,  verify that  the  cell  contamina-
 tion is  less than  10  ug/L  Cr by analyzing deminerallzed water and  the  appro-
 priate volume of  supporting  electrolyte   in  a  manner similar  to the procedure
 described  in 7.4.3  and  7.4.4.

     7.6   Calibration;

           7.6.1  After  running a differential   pulse   polarogram on  the  sample
     solution, quantitate   the   chromium  using  the  technique of  standard
     addition.
                                   7198 - 5
                                                          Revision       0
                                                          Date   September  1986

-------
2.0-
1.5-
                Current x 10^ nA
                26 Oct 81  No. 1
                Standard No. 1 DPP
                Initial E: 0.000 V
                Final E: -0.350 V

                Peak 1: -0.160V
                       1.181E2nA
                       250.0 ppb
                                                  2.0-
                                                  1.5-
       Current x 10^ n A
       26 OCT 81  No. 1
       Standard No. 2 DPP
       Initial E: 0.000 V
       Final E: -0.350 V

       Peak 1:-0.154 V
              1.146E3nA
              2.500 ppm
     -0.1    -0.3
-0.1   -0.3
     Figure 2. Typical differential pulse polarogram at 0.25 ppm and 2.5 ppm Cr
in 0.1 N buffer.
                             7198 -  6
                                                         Revision       0
                                                         Date   September  1986

-------
          7.6.2  Three standard  additions  should  be  made  to obtain the best
     precision and accuracy.   The  first   standard  addition should be approxi-
     mately one-half the concentration  of  the  sample, the second  equal  to that
     of the sample,  and the  third  about   1.5 times the sample concentration.
     The total volume due  to  standard additions  should not exceed the cell
     value by more than 10%.

          7.6.3  Add an appropriate aliquot of  chromium standard solution I,
     II, or III to the sample 1n  the  cell.     Deaerate for  30 sec  to  mix  the
     solution and remove oxygen added with the known addition.

          7.6.4  Repeat the analysis procedure,  beginning with Step 7.4.4  for
     each standard addition.

7.7  Calculations;

          7.7.1  Calculate  the concentration   of chromium  determined  by each
     standard addition procedure as follows:
                         c
                            -
                                            .
                          i - tjv, + (tr i,)v    vu


where:

     ij = Current peak height for the sample (nA);

     1j = Current peak height for the sample plus standard (nA);

     Vu = Volume of sample in the cell (ml);

     Vj = Volume of standard taken for spiking (ml);

      V = Volume in cell prior to standard addition;

     Cs = Concentration of standard used to spike (mg/L); and

     Cu = Concentration of the unknown in the sample (mg/L).

          7.7.2  Some  microprocessor   polarographic   systems  will  perform
     these calculations automatically.


8.0  QUALITY CONTROL

     8.1  All quality control data should be maintained and available for easy
reference or inspection.

     8.2  If necessary, dilute samples  so  that  they fall within the working
range.
                                  7198 - 7
                                                         Revision
                                                         Date  September 1986

-------
     8.3  Quant1 tat1on must be performed  by  the method of standard additions
(see Method 7000, Section 8.7).

     8.4  Verify calibration  with  an  Independently  prepared check standard
every 15 samples (see Chapter One, Section 1.1.8).

     8.5  Standards should be compared  to  a  reference standard on a routine
basis.
9.0  METHOD PERFORMANCE

     9.1  Precision and  accuracy  data  for  this  method  are  summarized 1n
Table 2.
10.0  REFERENCES

1.   Methods for  Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December 1982, Method 218.4 and 218.5.
                                   7198 - 8
                                                          Revision
                                                          Date  September 1986

-------
     TABLE 2.   PRECISION AND ACCURACY OF THE DPP OF HEXAVALENT CHROMIUM
2a.  Precision
       Sample type    No. of replicates
                        Average  value
                              % RSD
       Leachatea
                            1.87
                              0.69
2b.  Accuracy  (spike recovery data)
                  Spike level   No. of     Average %
     Sample type     (mg/L)     samples    recovery
                                        Standard
                                      deviation of
                                       % recovery
     EP extracts
5.0
8
92.8
6.4
2c.  Methods comparison
                      D1ff. pulse
                      polarography
                APDC extrac-
               tion ICAP-OES
                   Ion chromatography
                  coupled to ICAP-OES
       Value3
   1.87
      1.84
                1.91
     al_eachate sample from a waste disposal site.
                                   7198 - 9
                                                          Revision       0
                                                          Date   September  1986

-------
                             METHOD ?i9a
        HEXAVALENT CHROMIUM  (DIFFERENTIAL PULSE POLAROGRAPH)
C
  7. 1
       Soak
   voltemmetrIc
 cells overnight
 7.4.31  Place
      lelactrode
     assembly In
       solution:
     outgas with
      nitrogen
  7.2
        fllnse
      electrode
 assembly before
    and between
 sample analyses
  7.3
 Engage electrode
assembly:  displace
  mercury droos:
 initiate voltage
  ramp;  obtain
   polargram
  Set instrument
 7.4.1
                            7.S
   of chromium
    standard
    solution:
    deaerate
                                   Prior
                                  to ana
 during analysis
    verify that
  eel 1  contamin.
 is < 10 ug/1 Cr
        Plpet
        sample
 with hexavalant
   chromium into
    voltammetrlc
        cell
                           7.6.1
                                                     7.6.4
       Repeat
      for each
      standard
      addition
 starting with
 section 7.4.4
                                  Run
    differential
       pulse
   polarogram on
 sample solution
 7.
  Add ammonlacal
   electrolyte:
  adjust  volume
                                                      7.7
  Calculate
concentration
 of chromium
 7.6.ll

     Ouantltate
  chromium using
   technique  of
   standard add.
                                                    f      Stop      J
                          7198 - 10
                                                     Revision       Q
                                                     Date   September 1986

-------
                                 METHOD 7200

                COBALT (ATOMIC ABSORPTION. DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 If Interferences are suspected.

     3.2  Excesses  of  other  transition  metals  may  slightly  depress  the
response of cobalt.  Matrix  matching  or  the method of standard additions 1s
recommended.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Cobalt hollow cathode lamp.
          4.2.2  Wavelength:  240.7 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  A1r.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:   Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:  Dissolve  1.000 g of cobalt metal  (analytical
     reagent grade) 1n 20 mL of  1:1  HNC"3   and  dilute to 1 liter with Type II
     water.  Chloride or nitrate salts   of  cobalt (II) may be used.  Although
     numerous hydrated forms exist, they are not recommended unless the exact
     composition of the compound 1s known.  Alternatively, procure  a certified
     standard from a supplier and verify by comparison with a second standard.
                                  7200 - 1
                                                         Revision
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used  as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  add  and  at  the same
     concentration  as  will  result  in  the  sample  to  be  analyzed  after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.

7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics  for an aqueous sample free of inter-
ferences are:

     Optimum concentration  range:  0.5-5 mg/L with a wavelength of 240.7 nm.
     Sensitivity:  0.2 mg/L.
     Detection  limit:  0.05 mg/L.

     9.2   In a   single   laboratory,   analysis  of  a mixed  Industrial-domestic
waste  effluent,  digested with Method  3010,   at concentrations of  0.2,  1, and 5
mg/L gave   standard   deviations   of   +0.013,  +0.01,  and +0.05,  respectively.
Recoveries  at these  levels  were  98%  and 97%,  respectively.

     9.3   For concentrations of  cobalt  below 0.1 mg/L, the furnace procedure
 (Method 7201) is recommended.


 10.0   REFERENCES

 1.   Methods  for  Chemical   Analysis  of  Water   and Wastes, EPA-600/4-82-055,
 December  1982,  Method 219.1.
                                   7200 - 2
                                                          Revision
                                                          Date   September  1986

-------
                 METHOD 7200

COBALT (ATOMIC  ABSORPTION. DIRECT ASPIRATION)
              C
                5.0
                    Prepare
                   standards
                7.1
              	1  For
                    sample
               preparation see
                  chapter 3.
                section  3.1.3
                7.2
                Analyze using
                 Method 7000.
                 Section 7.2
              (      Stop       J
            7200 -  3
                                      Revision       0
                                      Date   September 1986

-------
                                 METHOD 7201

                COBALT (ATOMIC ABSORPTION.  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if Interferences are suspected.

     3.2  Excess chloride  may  Interfere.     It  is  necessary  to  verify by
standard additions that the interference is absent.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section  4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Drying time and temp:  30  sec at 125*C.
          4.2.2  Ashing time and temp:  30  sec at 900*C.
          4.2.3  Atomizing time and temp:  10 sec at 2700*C.
          4.2.4  Purge gas:  Argon.
          4.2.5  Wavelength:  240.7 nm.
          4.2.6  Background correction:  Required.
          4.2.7  Other operating parameters should be  set as specified by the
     particular instrument manufacturer.
     NOTE:  The above concentration values and instrument conditions are for a
            Perkin-Elmer HGA-2100, based  on  the  use  of  a 20-uL injection,
            continuous-flow purge  gas,  and  nonpyrolytic  graphite.  Smaller
            sizes of  furnace  devices  or  those  employing  faster  rates of
            atomization can be  operated  using lower atomization temperatures
            for shorter time periods than the above-recommended settings.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.
                                  7201 - 1
                                                         Revision
                                                         Date  September 1986

-------
     5.2  Preparation  of standards;

          5.2.1   Stock solution:   Dissolve  1.000 g of cobalt metal  (analytical
     reagent grade)  1n 20 ml of 1:1   HMOs   and  dilute to  1 liter with Type  II
     water.   Chloride  or nitrate salts  of   cobalt  (II) may be  used.  Although
     numerous hydrated forms exist,  they are not recommended unless  the  exact
     composition of the compound 1s  known.   Alternatively, procure  a  certified
     standard from a supplier and verify by comparison with a second  standard.

          5.2.2  Prepare  dilutions   of the stock   solution   to   be used  as
     calibration standards at the time of analysis.   The calibration  standards
     should  be  prepared  using  the  same  type   of add  and  at  the same
     concentrations as 1n the sample after  processing (0.5% v/v
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3,  Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.3, Furnace Procedure.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  Precision and accuracy data are not available at this time.

     9.2  The  performance  characteristics  for  an  aqueous  sample  free of
Interferences are:

     Optimum concentration  range:  5-100 ug/L.
     Detection limit:  1  ug/L.


10.0  REFERENCES

1.   Methods for   Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December  1982, Method  219.2.
                                   7201 - 2
                                                          Revision
                                                         Date  September  1986

-------
                METHOD 7ZO1

COBALT (ATOMIC  ABSORPTION.  FURNACE TECHNIQUE)
               5.O
                   Prepare
                  standards
               7. 1
                     For
                    sample
              preparation  see
                  chapter  3.
                 section 3.2
               7.2
               Analyze  using
                Method  7000.
                Section 7.3.
             f      Stop       J
           7201  - 3
                                     Revision       Q
                                     Date  September  1986

-------
                                 METHOD 7210

                COPPER (ATOMIC ABSORPTION.  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000.

     3.2  Background correction may be required because nonspecific absorption
and scattering can be  significant  at  the analytical wavelength.  Background
correction with certain Instruments may be difficult at this wavelength due to
low-intensity output  from  hydrogen  or  deuterium  lamps.   Consult specific
instrument manufacturer's literature for details.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Copper hollow cathode lamp.
          4.2.2  Wavelength:  324.7 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  A1r.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Recommended, if possible.


5.0  REAGENTS

     5.1  See Section 5.0 of  Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:    Dissolve  1.00  g  of  electrolytic copper
      (analytical reagent grade)  in  5 mL of  redistilled  HNOs   and dilute to
      1  liter with Type II water.   Alternatively, procure  a certified standard
      from a  supplier  and verify  by comparison with a  second standard.
                                   7210 - 1
                                                         Revision
                                                         Date  September  1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used as
     calibration standards at the time of analysis.  The  calibration standards
     should be prepared using the same type of add and at  the same concentra-
     tion as will result In the sample to be analyzed after processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and  Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the  sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics for an aqueous sample free of Inter-
ferences are:

     Optimum concentration range:  0.2-5 mg/L with a wavelength of 324.7 nm.
     Sensitivity:  0.1 mg/L.
     Detection  limit:  0.02 mg/L.

     9.2  Precision and accuracy data  are available 1n Method 220.1 of Methods
for Chemical Analysis of Water and Wastes.


10.0   REFERENCES

1.   Methods  for  Chemical  Analysis   of  Water  and Wastes, EPA-600/4-82-055,
December  1982,  Method  220.1.
                                   7210 - 2
                                                          Revision
                                                          Date  September 1986

-------
               METHOD 721O

COPPER  (ATOMIC ABSORPTION. OIHECT ASPIRATION)
             C


5.O 1
Prepare
standards

7.1
Si
preparat
Chaf
sect!

7.2

'or
imple
.Ion see
iter 3.
ion 3.2


Analyze using
Method 7000.
Section 7.2


            (     stop      j
             7210 - 3
                                     Revision       0
                                     Date  September  1986

-------
                                  METHOD  7211
                 COPPER  (ATOMIC  ABSORPTION,  FURNACE TECHNIQUE)
1.0  SCOPE AND APPLICATION
      1.1   See Section 1.0 of Method 7000.
2.0  SUMMARY OF METHOD
      2.1   See Section 2.0 of Method 7000.

3.0  INTERFERENCES
      3.1   If interferences are suspected, see Section 3.0 of Method 7000.
      3.2   Background correction may be required since nonspecific absorption
and scattering  can be significant  at  the analytical wavelength.   Background
correction with certain instruments may be difficult at this wavelength due to
low  intensity output  from hydrogen  or  deuterium  lamps.   Consult  specific
instrument manufacturer's literature for details.

4.0  APPARATUS AND MATERIALS
      4.1   For basic apparatus, see Section 4.0 of Method 7000.
      4.2   Instrument parameters (general):
            4.2.1  Drying time and temp:  30 sec at 125°C.
            4.2.2  Ashing time and temp:  30 sec at 900°C.
            4.2.3  Atomizing time and temp:  10 sec at 2700°C.
            4.2.4  Purge gas:   Argon or nitrogen.
            4.2.5  Wavelength:  324.7 nm.
            4.2.6  Background correction:  Recommended.
            4.2.7   Other operating parameters  should  be set  as  specified by
      the particular instrument manufacturer.
      NOTE; The above concentration values and instrument conditions are for a
            Perkin-Elmer  HGA-2100,  based  on the  use of  a 20-uL  injection,
            continuous-flow purge gas,  and nonpyrolytic graphite.  Smaller size
            furnace devices or those employing faster rates of atomization can
            be operated  using lower  atomization  temperatures  for  shorter time
            periods than the above-recommended settings.
                                   7211 - 1                       Revision 0
                                                                  July 1992

-------
5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards

            5.2.1   Stock  solution  - Dissolve  1.00  g of  electrolytic  copper
      (analytical regent grade) in 5 ml redistilled HNO, and dilute to 1 liter
      with water.  Alternatively,  procure a certified standard from a supplier
      and verify by comparison with a second standard.

            5.2.2   Prepare  dilutions  of  the  stock  solution to  be used  as
      calibration  standards  at  the  time  of  analysis.     The  calibration
      standards should be prepared using the same type of acid and at the same
      concentrations as  in the sample  to  be analyzed after  processing (0.5%
      v/v HN03).


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See Chapter Three, Step 3.1.3,  Sample Handling and Preservation.


7.0  PROCEDURE

      7.1   Sample Preparation - The procedures  for  preparation  of the sample
are given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.3,  Furnace Technique.


8.0  QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are not available at this time.

      9.2   The  performance  characteristics  for  an  aqueous  sample free  of
interferences are:

      Optimum concentration range:  5-100 ug/L.
      Detection limit:  1 ug/L.
                                   7211 - 2                       Revision 0
                                                                  July 1992

-------
10.0  REFERENCES

1.    Methods for  Chemical  Analysis of  Water and Wastes;  U.S.  Environmental
Protection  Agency.   Office  of   Research  and   Development.   Environmental
Monitoring  and  Support  Laboratory. ORD  Publication  Offices  of Center  for
Environmental  Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.
                                   7211 - 3                       Revision 0
                                                                  July 1992

-------
                    METHOD  7211
COPPER (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)
                     Start
                   S.0 Pnpar*
                    •tandard*
                 7.1 For «ampl«
                 preparation •••
                Chapter 3, Section
                      3 2
                                                                              i
                 7.2 Analyi* uung
                  M.thod 7000
                   Section 7.3
                     Stop
                       7211 -  4
Revision 0
July  1992

-------
                                 METHOD 7380
                 IRON (ATOMIC ABSORPTION,  DIRECT ASPIRATION)

1.0  SCOPE AND APPLICATION
     1.1  See Section 1.0 of Method 7000.

2.0  SUMMARY OF METHOD
     2.1  See Section 2.0 of Method 7000.

3.0  INTERFERENCES
     3.1  See Section 3.0 of Method 7000 If interferences are suspected.
     3.2  Iron 1s a universal contaminant,  and  great care should be taken to
avoid contamination.
4.0  APPARATUS AND MATERIALS
     4.1  For basic apparatus, see Section 4.0 of Method 7000.
     4.2  Instrument parameters  (general):
          4.2.1  Iron hollow cathode lamp.
          4.2.2  Wavelength:  248.3 nm  (primary); 248.8, 271.9, 302.1, 252.7,
                 or 372.0 nm (alternates).
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Required.
5.0  REAGENTS
     5.1  See Section 5.0 of Method 7000.
     5.2  Preparation of standards;
          5.2.1  Stock solution:    Dissolve  1.000  g  iron  wire (analytical
     reagent grade) in 10 mL redistilled HN03  and Type II water and dilute to
     1 liter with Type II  water.    Note that iron passivates in concentrated
     HN03, and thus some water   should  be  present.  Alternatively, procure a
     certified standard from a supplier and verify by comparison with a second
     standard.
                                  7380 - 1
                                                         Revision
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock   solution   to  be   used as
     calibration standards at the time of analysis.   The  calibration standards
     should  be  prepared  using  the  same  type  of  add  and   at   the  same
     concentration  as  will   result  1n  the  sample  to be   analyzed after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3, Sample Handling  and Preservation.

7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for   preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method  7000.


9.0  METHOD  PERFORMANCE

     9.1  The performance characteristics  for an aqueous sample free of inter-
ferences  are:

     Optimum concentration range:   0.3-5 mg/L with a wavelength of 248.3 nm.
     Sensitivity:   0.12  mg/L.
     Detection  limit:  0.03 mg/L.

     9.2   Precision and  accuracy data  are  available  In Method 236.1 of Methods
for  Chemical Analysis  of Water  and Wastes.


10.0  REFERENCES

1.   Methods for  Chemical   Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December  1982,  Method  236.1.
                                   7380 - 2
                                                          Revision
                                                          Date   September 1986

-------
               METHOD 7380

IRON (ATOMIC  ABSORPTION.  DIRECT ASPIRATION)
              5.O
                  Prepare
                 standards
              7.1
                  1  For
                   sample
             preparation 'sec
                 chapter 3.
                section 3.2
              Analyze using
               Method 7000.
               Section 7.2
            f      Stop       J
            7380 -  3
                                      Revision       0
                                      Date   September 1986

-------
                                 METHOD  7381

                  IRON  (ATOMIC  ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

      1.1   See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

      2.1   See Section 2.0 of Method 7000.


3.0  INTERFERENCES

      3.1   See Section 3.0 of Method 7000.

      3.2   Iron  is  a  universal contaminant,  particularly at the  low levels
determined by this method.  Great care should be taken to avoid contamination.


4.0  APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument parameters (general):

            4.2.1   Drying time and temp:  30 sec at 125°C.

            4.2.2   Ashing time and temp:  30 sec at 1000°C.

            4.2.3   Atomizing time and temp:  10 sec at 2700°C.

            4.2.4   Purge gas: Argon or nitrogen.

            4.2.5   Wavelength:  248.3 nm.

            4.2.6   Background Correction:  Recommended.

            4.2.7   Other operating parameters should be set as specified by the
      particular instrument manufacturer.

      NOTE: The above concentration values and instrument conditions are for a
            Perkin-Elmer  HGA-2100,  based  on  the  use  of  a  20-uL  injection,
            continuous-flow purge gas,  and nonpyrolytic  graphite.  Smaller size
            furnace devices or those employing faster rates of atomization can
            be operated using  lower  atomization  temperatures  for shorter time
            periods than the above-recommended settings.
                                   7381 - 1                       Revision 0
                                                                  July 1992

-------
5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards

            5.2.1   Stock  solution  - Dissolve  1.000  g iron wire  (analytical
      reagent grade) in 10 ml redistilled HN03 and water and dilute to 1 liter
      with water.  Note that iron passivates in concentrated HNO, and, thus, some
      water should be present.  Alternatively, procure  a certified standard from
      a supplier and verify by comparison with a second standard.

            5.2.2   Prepare  dilutions  of the stock  solution  to  be used  as
      calibration standards at the time of analysis.  The calibration standards
      should  be  prepared  using  the  same  type  of   acid   and  at  the  same
      concentrations as in the sample to  be analyzed after processing (0.5% v/v
      HN03).


6.0  SAMPLE COLLECTION, PRESERVATION AND HANDLING

      6.1   See Chapter Three, Step 3.1.3,  Sample Handling and Preservation.


7.0  PROCEDURE

      7.1   Sample preparation - The procedures for preparation of the sample are
given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.3,  Furnace Technique.


8.0  QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

      9.1   Precision and  accuracy data are not available at this time.

      9.2   The  performance  characteristics  for  an   aqueous  sample free  of
interferences are:

      Optimum concentration range:   5-100 ug/L.
      Detection limit:  1  ug/L.
                                   7381 - 2                       Revision 0
                                                                  July 1992

-------
10.0  REFERENCES

1.    Methods for  Chemical  Analysis of  Water  and Wastes;  U.S.  Environmental
Protection Agency.  Office of Research and Development. Environmental Monitoring
and Support  Laboratory.  ORD Publication Offices  of Center  for  Environmental
Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.
                                   7381 - 3                       Revision 0
                                                                  July 1992

-------
                  METHOD  7381
IRON  (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)
                     Start
                   5.0 Prepare
                    (tandard*
                 7.1 For *ampl«
                 preparation »••
               Chapter 3, Section
                      3.2
                7.2 Analyze u«ing
                  Method  7000
                  Section 7.3
                                                                              i
                     Stop
                     7381 -  4
Revision  0
July 1992

-------
                                 METHOD 7420

                 LEAD (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 If Interferences are suspected.

     3.2  Background correction Is required at either wavelength.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Lead hollow cathode lamp.
          4.2.2  Wavelength:  283.3 nm  (primary); 217.0 nm (alternate).
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  A1r.
          4.2.5  Type of  flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of  standards;

          5.2.1  Stock solution:  Dissolve  1.599  g of lead nitrate,
     (analytical  reagent grade),  in  Type  II  water,  acidify  with  10 mL
     redistilled HN03, and dilute to  1  liter  with  Type II water.  Alterna-
     tively, procure  a   certified  standard  from  a  supplier  and verify by
     comparison with a second standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared using  the  same  type  of  acid  and  at  the same
     concentration  as  will  result  in  the  sample  to  be  analyzed  after
     processing.

                                  7420 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING
     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.

7.0  PROCEDURE
     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given In Chapter Three, Section 3.2.
     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.

8.0  QUALITY CONTROL
     8.1  See Section 8.0 of Method 7000.

9.0  METHOD PERFORMANCE
     9.1  The  performance  characteristics  for  an  aqueous  sample  free of
Interferences are:
     Optimum concentration  range:  1-20 mg/L with a wavelength of 283.3 nm.
     Sensitivity:  0.5 mg/L.
     Detection limit:  0.1  mg/L.
     9.2  For concentrations of   lead  below  0.2  mg/L, the furnace technique
 (Method 7421) 1s  recommended.
     9.3  Precision  and  accuracy  data  are  available 1n Method  239.1 of Methods
 for  Chemical Analysis  of Water and Wastes.
     9.4  The data shown In Table  1  were obtained  from  records of state and
 contractor  laboratories.  The  data  are  Intended  to  show  the  precision of the
 combined  sample  preparation and analysis method.
 10.0  REFERENCES
 1.  Methods for  Chemical   Analysis  of  Water  and  Wastes,  EPA-600/4-82-055,
 December  1982, Method  239.1.
 2.  Gaskill,  A., Compilation  and Evaluation   of  RCRA Method  Performance  Data,
 Work Assignment  No.  2,  EPA Contract No.  68-01-7075, September 1986.
                                   7420 - 2
                                                          Revision
                                                          Date  September 1986

-------
                      TABLE 1. METHOD PERFORMANCE DATA
       Sample                      Preparation                Laboratory
       Matrix                        Method                   Replicates


Wastewater treatment sludge           3050                450, 404 ug/g

Emission control dust                 3050                42,500, 63,600 ug/g
                                  7420 - 3
                                                         Revision
                                                         Date  September 1986

-------
                METHOD 742O

LEAD (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
              s.o
                  Prepare
                 standards
              7. 1
                    For
                   sample
             preparation see
                 chapter. 3.
                section 3.2
              7.2
              Analyze  using
               Method  7000.
               Section 7.2
            f      Stop       J
          7420 - 4
                                    Revision       0
                                    Date  September  1986

-------
                                 METHOD 7421

                 LEAD (ATOMIC ABSORPTION.  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 1f Interferences are suspected.

     3.2  Background correction 1s required.

     3.3  If poor recoveries are obtained, a matrix modifier may be necessary.
Add 10 uL of phosphoric acid (Paragraph 5.3)  to 1 ml of prepared sample 1n the
furnace sampler cup and mix well.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Drying time and temp:  30*sec at 125*C.
          4.2.2  Ashing time and temp:  30*sec at 500'C.
          4.2.3  Atomizing time and temp:   10 sec at 2700*C.
          4.2.4  Purge gas:  Argon.
          4.2.5  Wavelength:  283.3 nm.
          4.2.6  Background correction:  Required.
          4.2.7  Other operating parameters should be  set as specified by the
                 particular Instrument manufacturer.
     NOTE:  The above concentration values and Instrument conditions are for a
            Perkln-Elmer HGA-2100, based  on  the  use  of  a 20-uL injection,
            continuous-flow purge  gas,  and  nonpyrolytlc  graphite.  Smaller
            sizes of  furnace  devices  or  those  employing  faster  rates of
            atomizatlon can be  operated  using lower atomlzation temperatures
            for shorter time periods than the above-recommended settings.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.
                                  7421 - 1
                                                         Revision
                                                         Date  September 1986

-------
     5.2   Preparation of  standards;

          5.2.1   Stock  solution:  Dissolve  1.599  g of lead nitrate, Pb(NOs)2
     (analytical   reagent grade),   In  Type   II  water,  acidify  with  10 ml
     redistilled  HN03,  and dilute to 1   liter with  Type II water.  Alterna-
     tively,  procure  a  certified   standard   from  a  supplier  and verify by
     comparison with  a  second  standard.

          5.2.2   Prepare  dilutions   of   the   stock  solution  to  be  used as
     calibration  standards at  the time of analysis.  The  calibration standards
     should  be   prepared using  the same  type  of  add  and   at  the  same
     concentrations as  in the  sample after processing  (0.5%  v/v HNOs).

     5.3   Phosphoric acid;   Reagent  grade.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1   See Chapter Three, Section 3.1.3, Sample  Handling  and  Preservation.

7.0  PROCEDURE

     7.1   Sample preparation;   The  procedures  for  preparation  of the sample
are given in Chapter Three,  Section 3.2.

     7.2  See Method 7000,  Paragraph 7.3,  Furnace Procedure.   The calculation
is given  in Method 7000, Paragraph 7.4.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD  PERFORMANCE

     9.1  Precision and  accuracy data are  available in Method 239.2 of Methods
for  Chemical Analysis of Water and Wastes.

     9.2  The  performance  characteristics   for  an  aqueous  sample  free of
interferences are:

     Optimum concentration  range:   5-100 ug/L.
     Detection limit:  1 ug/L.

     9.3  The data shown in Table   1  were  obtained from records of state and
contractor laboratories.  The data  are   intended  to show the precision of the
combined  sample  preparation and analysis method.
                                   7421 - 2
                                                          Revision
                                                          Date   September  1986

-------
10.0  REFERENCES
1.   Lead by Flameless Atomic  Absorption  with Phosphate Matrix Modification,
Atomic Spectroscopy, !_ (1980), no. 3, pp. 80-81.
2.   Gasklll, A., Compilation and Evaluation  of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075, September 1986.
                                  7421 - 3
                                                         Revision
                                                         Date  September  1986

-------
                      TABLE 1.  METHOD PERFORMANCE DATA
Sample
Matrix
Contaminated soil
Paint primer
Lagoon soil
NBS SRM 1646 Estuarlne sediment
NBS SRM 1085 Wear metals 1n
lubricating oil
Solvent extracted oily waste
Preparation
Method
3050
3050
3050
3050
3030
3030
Laboratory
Replicates
163, 120 mg/g
0.55, 0.63 mg/g
10.1, 10.0 ug/g
23.7 ug/g*
274, 298 ug/gb
9, 18 ug/L
aB1as of -16% from expected.

&B1as of -10 and -2% from expected, respectively,
                                   7421 - 4
                                                         Revision      0
                                                         Date  September  1986

-------
               METHOD 7431

LEAD (ATOMIC ABSORPTION.  FURNAC6 TECHNIQUE)
           C
             5.0
                 Prepare
                standards
7. 1
prepar
ct
sec
For
sample
•ation see
lapter 3.
tion 3.2
             7.2
             Analyze using
              Method 7000.
              Section 7.3.
            calculation  7.4
           (      Stop       J
            7421 -  5
                                      Revision       0
                                      Date  September  1986

-------
                                 METHOD  7430
                LITHIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)

1.0   SCOPE AND APPLICATION
      1.1   See Section 1.0 of Method 7000.

2.0   SUMMARY OF METHOD
      2.1   See Section 2.0 of Method 7000.

3.0   INTERFERENCES
      3.1   See Section 3.0 of Method 7000 if interferences are suspected.

4.0   APPARATUS AND MATERIALS
      4.1   For basic apparatus, see Section 4.0 of Method 7000.
      4.2   Instrument parameters (general):
            4.2.1   Lithium hollow cathode lamp.
            4.2.2   Wavelength:  670.8 nm.
            4.2.3   Fuel:  Acetylene.
            4.2.4   Oxidant:  Air.
            4.2.5   Type of flame:   Oxidizing (fuel lean).
            4.2.6   Background Correction:  Not required.

5.0   REAGENTS
      5.1   See Section 5.0 of Method 7000.
      5.2   Preparation of standards
            5.2.1   Stock  solution:   (1.0 mL =  1.0 mg  Li).   Dissolve  5.324 g
      lithium carbonate, Li,C03,  in a minimum volume of  1:1  HC1  and dilute to
      1 liter with water.  Alternatively, procure a certified standard  from a
      supplier and verify by comparison with a second standard.

                                   7430  - 1                       Revision 0
                                                                  July 1992

-------
            5.2.2   Prepare  dilutions  of  the  stock  solution  to  be used  as
      calibration standards at the  time of  analysis.  The calibration standards
      should be  prepared  using the same type of  acid as the samples  used  to
      prepare the samples  and cover the range of expected concentrations in the
      samples.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See Chapter Three, Step 3.1.3, Sample Handling and Preservation.


7.0   PROCEDURE

      7.1   Sample preparation - The procedures for preparation of the sample are
given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.2,  Direct Aspiration.


8.0   QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.


9.0   METHOD PERFORMANCE

      9.1   The  performance  characteristics  for  an  aqueous  sample free  of
interferences are:

      Optimum concentration range:  0.1-2 mg/L at a wavelength of 670.8 nm.
      Sensitivity: 0.04 mg/L.
      Detection limit: 0.002 mg/L.


10.0 REFERENCES

1.    Standard Methods  for the Examination of Water and Wastewater, 16th ed.;
      Greenberg,  A.E.;  Trussell,  R.R.;  Clesceri,  L.S.,  Eds.;  American Water
      Works  Association,  Water Pollution  Control  Federation,  American Public
      Health Association:  Washington,  DC, 1985.
                                   7430 - 2                       Revision 0
                                                                  July 1992

-------
                    METHOD 7430
LITHIUM (ATOMIC ABSORPTION, DIRECT ASPIRATION)
                        Start
                      S.0 Prepare
                      •tandardi
                    7.1 For »ample
                   preparation, •••
                  Chapter 3,  Step 3.2
                   7 2 Analyze using
                   Method 7000, Step
                         7.2
                        Stop
                      7430 - 3
Revision 0
July  1992

-------
                                 METHOD 7450

              MAGNESIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  All elements forming stable oxyanions (P,  B,  Si, Cr, S, V, Ti, Al,
etc.) will complex magnesium and  interfere  unless  lanthanum is added.  (See
Method 7000, Paragraph  3.1.1.)    Addition  of  lanthanum to prepared samples
rarely presents a problem because  virtually all environmental  samples contain
sufficient magnesium to require  dilution  to  be  in  the linear range of the
method.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Magnesium hollow cathode lamp.
          4.2.2  Wavelength:   285.2 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of  flame:  Oxidizing  (fuel  lean).
          4.2.6  Background  correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of  standards;

          5.2.1  Stock   solution:     Dissolve   1.000  g  of  magnesium  metal
     (analytical reagent  grade)  in 20 mL 1:1   HNOs and dilute  to  1  liter with
     Type II  water.    Alternatively,   procure   a  certified   standard from  a
     supplier and verify  by  comparison with a  second standard.


                                  7450 - 1
                                                         Revision       0
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions   of  the   stock   solution  to  be  used as
     calibration standards at the time of analysis.   The calibration standards
     should  be  prepared  using  the  same   type  of  add  and  at  the same
     concentration  as  will   result  1n  the  sample  to  be  analyzed  after
     processing, Including 1  ml  lanthanum  solution   per  10 ml solution (see
     Paragraph 3.2).

          5.2.3  Lanthanum chloride solution:  Dissolve 29  g 13203 in  250 ml
     concentrated HC1 -
          (CAUTION:  REACTION IS VIOLENT!) -
     and dilute to 500 ml with Type II water.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3,  Sample Handling  and  Preservation.

7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the  sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The   performance  characteristics  for  an  aqueous  sample   free of
Interferences are:

     Optimum concentration range:  0.02-0.05 mg/L with  a wavelength  of 285.2
                                   nm.
     Sensitivity:   0.007  mg/L.
     Detection  limit:   0.001 mg/L.

     9.2   In a   single   laboratory,   analysis  of  a mixed  industrial-domestic
waste effluent,  digested  with Method   3010,  at  concentrations of 2.1 and 8.2
mg/L gave  standard deviations of +0.1   and  +0.2, respectively.  Recoveries at
both of these  levels were 100%.
 10.0   REFERENCES

 1.    Methods  for  Chemical   Analysis   of  Water  and Wastes, EPA-600/4-82-055,
 December 1982,  Method  242.1.
                                   7450 -  2
                                                         Revision
                                                         Date  September 1986

-------
                   METHOD 7450

MAGNESIUM (ATOMIC  ABSORPTION.  DIRECT ASPIRATION)
                     Prepare
                    standards
                 7.,
                  . - 1 For
                      sample
                preparation see
                    chapter 3.
                   section 3.2
                 7.2
                 Analyze using
                  Method 7000.
                  Section 7.2
               f     StOP      J
             7450 - 3
                                       Revision       0
                                       Date   September 1986

-------
                                 METHOD 7460

              MANGANESE (ATOMIC ABSORPTION, DIRECT  ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction is required.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Manganese hollow cathode lamp.
          4.2.2  Wavelength:  279.5 nm (primary); 403.1 nm (alternate).
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxidant:  Air.
          4.2.5  Type of flame:    Slightly  oxidizing  (slightly fuel-lean to
                 stoichiometric).
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards:

          5.2.1  Stock solution:  Dissolve 1.000 g manganese metal (analytical
     reagent grade)  in 10 mL  redistilled  HN03  and dilute to  1 liter  with
     Type II  water.    Alternatively,   procure  a  certified  standard from a
     supplier and verify by comparison with a  second standard.

          5.2.2  Prepare  dilutions  of  the   stock  solution  to  be  used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  add  and  at  the same
     concentration   as  will  result   1n the  sample  to  be  analyzed  after
     processing.

                                  7460 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics  for an aqueous sample free of inter-
ferences  are:

     Optimum concentration  range:  0.1-3 mg/L with a wavelength of 279.5 nm.
     Sensitivity:   0.05 mg/L.
     Detection  limit:  0.01 mg/L.

     9.2  Precision and accuracy data  are  available 1n Method 243.1 of Methods
for  Chemical Analysis  of Water  and Wastes.


10.0  REFERENCES

1.   Methods for   Chemical  Analysis   of   Water  and Wastes, EPA-600/4-82-055,
December  1982,  Method  243.1.
                                   7460 - 2
                                                          Revision
                                                          Date  September 1986

-------
                  METHOD 7460

MANGANESE  (ATOMIC ABSORPTION. DIRECT ASPIRATION)
5
0 1
Prepare
standards
7. 1
prepar
cf
set
For
sample
-ation see
lapter 3.
tlon 3.2
                7 .Z
                Analyze using
                 Method 7OOO.
                 Section 7.3
              f     Stop      J
           7460 - 3
                                     Revision       0
                                     Date  September 1986

-------
                                 METHOD  7461

               MANGANESE (ATOMIC ABSORPTION. FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

      1.1   See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

      2.1   See Section 2.0 of Method 7000.


3.0  INTERFERENCES

      3.1   See Section 3.0 of Method 7000.

      3.2   Background correction must be used.


4.0  APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument parameters (general):

            4.2.1   Drying time and temp:  30 sec at 125°C.

            4.2.2   Ashing time and temp:  30 sec at 1000°C.

            4.2.3   Atomizing time and temp:  10 sec at 2700°C.

            4.2.4   Purge gas:  Argon or nitrogen.

            4.2.5   Wavelength:  279.5 nm.

            4.2.6   Background correction:  Required.

            4.2.7   Other operating parameters  should be set as specified by the
      particular instrument manufacturer.

      NOTE: The above concentration values and instrument conditions are for a
            Perkin-Elmer  HGA-2100,  based  on  the  use  of  a  20  uL  injection,
            continuous-flow purge gas,  and nonpyrolytic graphite.  Smaller size
            furnace devices or those employing faster rates of atomization can
            be operated using  lower atomization  temperatures  for shorter time
            periods than the above-recommended settings.
                                   7461 - 1                       Revision 0
                                                                  July 1992

-------
5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards

            5.2.1   Stock  solution  -  Dissolve  1.000  g  manganese    metal
      (analytical reagent grade)  in  10 ml redistilled HN03 and dilute to 1 liter
      with water.  Alternatively, procure a certified standard from a supplier
      and verify by comparison with a second standard.

            5.2.2   Prepare  dilutions  of  the  stock  solution  to  be used  as
      calibration standards at the time of analysis.  The calibrations standards
      should  be  prepared  using the  same  type  of  acid  and  at  the  same
      concentrations as in the sample after processing (0.5% v/v HN03).


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See Chapter Three, Step 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

      7.1   Sample Preparation - The procedures  for preparation of the sample are
given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.3, Furnace Technique.


8.0  QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are not available at this time.

      9.2   The  performance  characteristics  for  an  aqueous  sample free  of
interferences are:

      Optimum concentration range:  1-30 ug/L.
      Detection limit:  0.2 ug/L.


10.0  REFERENCES

1.    Methods  for  Chemical Analysis  of  Water  and  Hastes;  U.S.  Environmental
Protection Agency. Office of Research and  Development. Environmental Monitoring
and  Support  Laboratory.  ORD  Publication  Offices of  Center  for Environmental
Research Information:  Cincinnati, OH, 1983; EPA-600/4-79-020.


                                   7461 - 2                       Revision 0
                                                                  July 1992

-------
                     METHOD  7461
MANGANESE  (ATOMIC  ABSORPTION,  FURNACE  TECHNIQUE)
                        Start
                      5.0 Prepare
                      «tandard«
                    7  1 For iample
                    preparation mm*
                  Chapter 3, Section
                         3.2
                   7.2 Analyze uiing
                     Method 7000
                     Section 7 3
                        Stop
                        7461  - 3
Revision 0
July 1992

-------
                                 METHOD 7470A

             MERCURY IN LIQUID WASTE (MANUAL COLD-VAPOR TECHNIQUE)
1.0   SCOPE AND APPLICATION

      1.1   Method 7470 is a  cold-vapor atomic absorption procedure approved for
determining the concentration of mercury in mobility-procedure extracts, aqueous
wastes, and ground waters.  (Method 7470  can also  be  used  for analyzing certain
solid and  sludge-type wastes;  however,  Method  7471  is usually  the method of
choice for these waste types.)  All samples must be subjected to  an  appropriate
dissolution step prior to  analysis.

2.0   SUMMARY OF METHOD

      2.1   Prior to analysis, the liquid samples  must  be  prepared  according to
the procedure discussed in this method.

      2.2   Method 7470, a cold-vapor atomic absorption technique,  is based on
the absorption of radiation at 253.7-nm by mercury  vapor.  The mercury is reduced
to the elemental  state and  aerated  from solution in a  closed system.  The mercury
vapor passes through a cell positioned in the light path of an atomic absorption
spectrophotometer.  Absorbance (peak height) is measured as a function of mercury
concentration.

      2.3   The typical detection  limit for this method is 0.0002 mg/L.

3.0   INTERFERENCES

      3.1   Potassium permanganate is added to eliminate possible  interference
from sulfide.   Concentrations as high as 20 mg/L of sulfide as sodium sulfide do
not interfere with the recovery of added inorganic mercury from reagent water.

      3.2   Copper has also been reported to interfere;  however, copper concen-
trations as high as  10 mg/L  had no effect  on recovery of mercury  from spiked
samples.

      3.3   Seawaters, brines,  and industrial effluents  high in chlorides require
additional  permanganate (as much as 25 mL) because, during the oxidation step,
chlorides are converted to free chlorine,  which also  absorbs  radiation of 253.7
nm.  Care must therefore be taken  to ensure that free chlorine is absent before
the mercury is reduced  and swept  into the cell.  This may be accomplished by
using an excess of hydroxylamine sulfate reagent (25 mL).  In  addition, the dead
air space in the BOD bottle must be purged before adding stannous  sulfate.  Both
inorganic and organic  mercury spikes  have  been quantitatively  recovered from
seawater by using this technique.

      3.4   Certain volatile  organic materials that absorb at this wavelength may
also cause  interference.  A preliminary run without reagents should determine if
this type of interference is  present.
                                  7470A  -  1                       Revision 1
                                                                  September 1994

-------
4.0   APPARATUS AND MATERIALS

      4.1   Atomic  absorption spectrophotometer  or equivalent:    Any  atomic
absorption unit  with  an open sample  presentation  area in which  to mount the
absorption cell is suitable.   Instrument  settings recommended by the particular
manufacturer  should  be followed.   Instruments designed  specifically  for the
measurement of mercury using the cold-vapor technique are commercially available
and may be substituted for the atomic absorption spectrophotometer.

      4.2   Mercury hollow cathode lamp or electrodeless discharge lamp.

      4.3   Recorder:  Any multirange variable-speed recorder that  is compatible
with the UV detection system  is suitable.

      4.4   Absorption cell:  Standard spectrophotometer cells 10 cm long with
quartz end windows may be used. Suitable cells may be constructed from Plexiglas
tubing,  1 in.  O.D.   x  4.5  in.    The ends  are ground  perpendicular  to the
longitudinal  axis, and quartz windows  (1  in.  diameter x 1/16 in. thickness) are
cemented in place.  The cell  is strapped  to a burner  for support and aligned in
the light beam by use of two 2-in.  x 2-in. cards.  One-in.-diameter  holes  are cut
in the middle of each  card.  The cards  are then placed over each  end of the cell.
The cell is then positioned and adjusted vertically and  horizontally to give the
maximum transmittance.

      4.5   Air pump:  Any peristaltic pump capable of delivering 1  liter air/min
may be used.   A Masterflex  pump with electronic  speed control has  been found to
be satisfactory.

      4.6   Flowmeter:  Capable of measuring an air flow of 1 liter/min.

      4.7   Aeration tubing:   A straight glass frit with a coarse porosity. Tygon
tubing  is  used  for passage of the mercury  vapor from the sample bottle to the
absorption cell and return.

      4.8   Drying tube:  6-in. x 3/4-in.-diameter  tube containing 20 g of mag-
nesium perchlorate or a small reading lamp with 60-W bulb which may be used to
prevent condensation of moisture  inside the cell.  The lamp should  be positioned
to shine on the absorption cell so  that the air temperature in the  cell is about
10°C above ambient.

      4.9   The  cold-vapor  generator  is  assembled  as  shown  in  Figure  1  of
reference  1  or  according  to the  instrument manufacturers  instructions.  The
apparatus  shown  in  Figure 1  is  a closed system.   An  open  system,  where the
mercury  vapor is passed through  the  absorption cell  only once,   may  be used
instead of the closed system.  Because mercury vapor is toxic,  precaution must
be taken to avoid its inhalation.   Therefore, a  bypass has  been included in the
system either to vent  the mercury vapor into an exhaust hood or to pass the vapor
through some absorbing medium, such as:

            1.   Equal volumes of 0.1 M KMn04 and 10% H2S04;  or

            2.  0.25% Iodine  in a  3% KI solution.


                                   7470A -  2                       Revision 1
                                                                  September 1994

-------
      A  specially treated  charcoal  that  will  adsorb  mercury vapor  is  also
      available  from Barnebey and Cheney,  East 8th Avenue  and North Cassidy
      Street, Columbus, Ohio 43219, Cat. #580-13 or #580-22.

      4.10   Hot  plate  or  equivalent  -  Adjustable  and capable of maintaining a
temperature  of 90-95°C.

      4.11   Graduated  cylinder or equivalent.

5.0   REAGENTS

      5.1    Reagent  Water:    Reagent  water  will  be  interference free.   All
references to water in this method will  refer to reagent water unless otherwise
specified.

      5.2    Sulfuric acid (H2S04),  concentrated:  Reagent  grade.

      5.3    Sulfuric acid, 0.5 N:  Dilute 14.0 ml of concentrated  sulfuric  acid
to 1.0 liter.

      5.4    Nitric  acid  (HN03),  concentrated:   Reagent grade  of low mercury
content.  If a high reagent blank is obtained,  it may be necessary to distill the
nitric acid.

      5.5    Stannous sulfate:   Add 25 g  stannous  sulfate to 250  ml  of 0.5 N
H2S04.   This mixture is a suspension  and should be stirred continuously during
use.  (Stannous chloride may be used in place of stannous sulfate.)

      5.6    Sodium chloride-hydroxylamine  sulfate  solution:   Dissolve 12 g of
sodium chloride and 12  g of hydroxylamine sulfate in reagent water  and dilute to
100 ml.   (Hydroxylamine hydrochloride may be  used in  place of hydroxylamine
sulfate.)

      5.7    Potassium permanganate, mercury-free,  5% solution  (w/v):  Dissolve
5 g of potassium permanganate in 100 mL of reagent water.

      5.8    Potassium persulfate, 5% solution (w/v):  Dissolve 5 g  of potassium
persulfate in 100 ml of reagent water.

      5.9    Stock mercury solution:  Dissolve 0.1354 g  of mercuric chloride in
75 ml of reagent  water. Add 10 ml of concentrated HN03  and adjust  the volume to
100.0 ml (1 ml = 1 mg Hg).  Stock solutions may also be purchased.

      5.10   Mercury working standard:  Make  successive dilutions of the stock
mercury solution to obtain a working standard containing  0.1 ug per ml.   This
working standard  and  the dilutions  of the  stock  mercury solution  should be
prepared fresh daily.  Acidity of the working standard  should be maintained at
0.15% nitric acid.  This acid should be  added to  the  flask,  as needed,  before
addition of the aliquot.
                                   7470A  -  3                       Revision 1
                                                                  September 1994

-------
6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must have been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with detergents, acids, and
reagent water.  Plastic and glass containers are both suitable.

      6.3   Aqueous  samples must  be acidified  to  a  pH  <2  with HN03.   The
suggested maximum holding times for mercury is 28 days.

      6.4   Nonaqueous samples shall be refrigerated, when possible, and analyzed
as soon as possible.

7.0   PROCEDURE

      7.1   Sample preparation:   Transfer  100 mL,  or  an   aliquot diluted  to
100 mL, containing <1.0 g of mercury, to a 300-mL BOD  bottle  or equivalent.  Add
5 mL of H2S04  and  2.5 mL of  concentrated  HN03, mixing after each addition.  Add
15 mL of potassium permanganate  solution  to each  sample bottle.  Sewage samples
may require additional permanganate.   Ensure that equal amounts of permanganate
are  added  to  standards  and  blanks.   Shake  and add  additional  portions  of
potassium permanganate solution, if necessary,  until  the purple color persists
for at least 15 min.  Add 8 mL of potassium persulfate to each bottle and heat
for 2  hr in a  water bath  maintained  at 95°C.   Cool  and  add 6  mL  of sodium
chloride-hydroxylamine sulfate to reduce  the excess permanganate.   After a delay
of at least 30 sec,  add 5 mL of  stannous  sulfate, immediately  attach the bottle
to the aeration apparatus,  and continue as described in Paragraph 7.3.

      7.2   Standard preparation:  Transfer 0-, 0.5-, 1.0-, 2.0-, 5.0-,  and 10.0-
mL aliquots of the mercury working standard, containing 0-1.0  ug of mercury, to
a series of 300-mL BOD bottles.  Add enough reagent water to  each bottle to make
a total volume of 100 mL.  Mix thoroughly  and add 5 mL of concentrated H2S04 and
2.5 mL of concentrated HN03  to each bottle.  Add 15 mL of KMn04 solution to each
bottle and allow to stand at least 15 min.  Add 8 mL of potassium persulfate to
each bottle and heat for 2 hr in a water  bath maintained at  95°C.   Cool and add
6 mL  of sodium chloride-hydroxylamine sulfate  solution  to reduce  the excess
permanganate.   When the solution has  been decolorized,  wait  30 sec, add 5 mL of
the stannous  sulfate solution,  immediately attach the bottle to  the aeration
apparatus, and continue as  described in Paragraph 7.3.

      7.3   Analysis:   At  this  point the sample is  allowed  to  stand quietly
without  manual  agitation.    The circulating pump,  which  has previously been
adjusted  to a  rate  of  1   liter/min,  is  allowed  to  run  continuously.   The
absorbance will  increase  and reach a maximum within  30  sec.   As  soon  as the
recorder pen levels  off (approximately 1 min), open the bypass valve and continue
the aeration until the absorbance returns  to its minimum value.   Close the bypass
valve,  remove  the  stopper  and  frit from  the  BOD  bottle,   and   continue  the
aeration.  Because of instrument  variation refer to the manufacturers recommended
operating conditions when using this method.
                                   7470A -  4                      Revision 1
                                                                  September 1994

-------
      7.4   Construct a calibration curve by plotting the absorbances of stan-
dards versus micrograms of mercury.   Determine  the  peak height of the unknown
from the chart and read the mercury value  from the standard curve.  Duplicates,
spiked samples, and check standards should be routinely analyzed.

      7.5   Calculate  metal   concentrations  (1)  by  the method of  standard
additions, or  (2) from  a  calibration curve.   All  dilution  or concentration
factors must be taken into account. Concentrations  reported for multiphased or
wet samples must be appropriately  qualified (e.g., 5 ug/g dry weight).

8.0   QUALITY CONTROL

      8.1  Refer to section 8.0 of Method 7000.

9.0   METHOD PERFORMANCE

      9.1   Precision and accuracy data are available in Method 245.1 of Methods
for Chemical Analysis of Water and Wastes.

10.0  REFERENCES

1.    Methods  for Chemical   Analysis  of  Water  and Wastes,  EPA-600/4-82-055,
December 1982, Method 245.1.
                                  7470A  - 5                       Revision 1
                                                                  September 1994

-------
                             METHOD 7470A
MERCURY IN  LIQUID WASTE  (MANUAL  COLD-VAPOR  TECHNIQUE)
              Sample Preparation
                   Standard Preparation
  7.1 Transfer aliquot
  to bottle, add H2S04
  and HNOi and mix.
    7.1  Add KMn04
      and shake.
                                7.2 Tranafer aliquot
                                 of the Hg working
                                   •tandard to
                                     bottle.
                                  7.2 Add raagent
                                  water, mix, add
                                   concentrated
                                  HjSC^and
                               7.1 Add more
                               permanganate
                                if neceeaary.
       7.1 Add
      potassium
    pereulfete, heat
    for 2 hrs., cool.
                                   7.2 Add KMn04
                                     potaaaium
                                  peraulfate, heat
                                 for 2 hrs. and cool.
                                   7.2 Add eodium
                                      chloride-
                                   hydroxylamine
                                   eulfate, wait 30
                                      eeconda.
    7.1  Add sodium
       chloride-
     hydroxylamine
    sulfate, wait 30
       seconds.
    7.1 Add atannoua
     sulfate, attach
       to aeretion
       apparatus.
 7.3 Analyze
   aample.
7.2 Add stannous
 aulfate, attach
   to aeration
   apparatue.
 7.4 Conatruct
   calibration
curve, determine
peak height and
   Hg value.
                                 7.4 Routinely
                               analyze duplicates,
                                spiked samples.
                                  7.5 Calculate
                                     metal
                                 concentrations.
                                (     Stop     J


                               7470A -  6
                                            Revision 1
                                            September 1994

-------
                                 METHOD 7471A

       MERCURY IN SOLID OR SEMISOLID WASTE  (MANUAL  COLD-VAPOR TECHNIQUE)


1.0   SCOPE AND APPLICATION

      1.1   Method 7471  is approved for measuring total  mercury (organic and
inorganic) in soils,  sediments,  bottom deposits,  and sludge-type materials. All
samples must be subjected to an  appropriate dissolution  step  prior to analysis.
If this dissolution  procedure is  not sufficient  to dissolve a specific matrix
type or sample, then this method is not applicable for that matrix.

2.0   SUMMARY OF METHOD

      2.1   Prior to analysis,  the solid or semi-solid samples must be prepared
according to the procedures discussed in this method.

      2.2   Method 7471, a cold-vapor atomic absorption  method, is based on the
absorption of radiation at the 253.7-nm wavelength by mercury vapor.  The mercury
is reduced to the elemental state  and aerated from  solution in a closed system.
The mercury vapor passes through a cell positioned in the light path  of an atomic
absorption spectrophotometer.  Absorbance (peak height) is measured as a function
of mercury concentration.

      2.3   The  typical  instrument detection  limit  (IDL) for this  method is
0.0002 mg/L.

3.0   INTERFERENCES

      3.1   Potassium permanganate is added to eliminate possible interference
from sulfide.   Concentrations as high as 20 mg/Kg  of sulfide,  as sodium sulfide,
do not interfere with the recovery of added inorganic mercury in reagent water.

      3.2   Copper has also been reported to interfere;  however, copper concen-
trations as high as  10 mg/Kg had  no  effect on recovery  of mercury  from spiked
samples.

      3.3   Samples high in chlorides require additional permanganate (as much
as 25 ml) because, during  the oxidation step,  chlorides are converted  to free
chlorine, which also  absorbs radiation of 253 nm.   Care must therefore be taken
to ensure that free chlorine is absent before the mercury is reduced and swept
into the cell.   This may be accomplished  by  using an excess of  hydroxylamine
sulfate reagent (25 ml).   In addition, the dead air  space in the BOD bottle must
be purged before adding stannous sulfate.

      3.4   Certain volatile organic materials that  absorb at this wavelength may
also cause interference.  A preliminary run  without  reagents should determine if
this type of interference is present.

4.0   APPARATUS AND MATERIALS

      4.1   Atomic absorption  spectrophotometer or  equivalent:    Any  atomic
absorption unit  with  an open sample  presentation  area  in which to mount  the

                                   7471A -  1                       Revision 1
                                                                  September  1994

-------
absorption cell is suitable.   Instrument  settings recommended by the particular
manufacturer  should  be followed.   Instruments designed  specifically  for the
measurement of mercury using the cold-vapor technique are commercially available
and may be substituted for the atomic absorption spectrophotometer.

      4.2   Mercury hollow cathode lamp or electrodeless discharge lamp.

      4.3   Recorder:  Any multirange variable-speed recorder that  is compatible
with the UV detection  system is suitable.

      4.4   Absorption cell:  Standard spectrophotometer cells 10 cm long with
quartz end windows may be used.  Suitable cells may be constructed from Plexiglas
tubing,  1  in.  O.D.   x 4.5  in.    The  ends  are ground  perpendicular  to the
longitudinal  axis, and quartz windows  (1  in.  diameter x 1/16 in. thickness) are
cemented in place.  The cell  is strapped  to  a burner  for support and aligned in
the light beam by  use of two 2-in.  x 2-in. cards.  One-in.-diameter  holes  are cut
in the middle  of each  card.  The cards are then placed over each  end of the cell.
The cell  is then positioned and adjusted vertically and  horizontally to give the
maximum transmittance.

      4.5   Air pump:   Any peristaltic pump capable of del ivering 1  L/min  air may
be used.   A Masterflex pump with electronic speed control  has been found to be
satisfactory.

      4.6   Flowmeter:  Capable of measuring an air flow of 1 L/min.

      4.7   Aeration tubing:  A straight glass frit with a coarse porosity. Tygon
tubing is used  for passage of the mercury  vapor from the  sample  bottle to the
absorption cell and return.

      4.8   Drying tube:   6-in.  x 3/4-in.-diameter tube  containing  20  g  of
magnesium perch!orate  or a small  reading lamp with 60-W bulb which may be used
to  prevent condensation  of  moisture  inside the cell.    The  lamp should  be
positioned to  shine on the absorption  cell  so  that  the air temperature in the
cell is about  10°C above  ambient.

      4.9   The cold-vapor  generator  is  assembled  as  shown  in  Figure  1  of
reference  1 or according to the  instrument manufacturers  instructions.   The
apparatus  shown in  Figure 1 is  a closed system.   An open  system, where the
mercury vapor  is  passed  through  the  absorption cell  only once,   may  be used
instead of the closed  system. Because mercury vapor is toxic, precaution must be
taken to avoid its  inhalation.   Therefore,  a bypass has  been  included  in the
system either  to  vent the  mercury vapor into  an exhaust  hood or to  pass the
vapor through  some absorbing medium,  such as:

            1.    equal volumes of 0.1  M KMn04  and  10% H2S04, or
            2.    0.25% iodine in a 3% KI solution.
      A  specially  treated  charcoal  that  will  adsorb  mercury  vapor  is  also
      available  from  Barneby and  Cheney,  East 8th  Avenue and  North Cassidy
      Street, Columbus, Ohio 43219, Cat. #580-13 or #580-22.
                                   7471A -  2                       Revision 1
                                                                  September 1994

-------
      4.10  Hot  plate  or equivalent  -  Adjustable and capable of maintaining a
temperature of 90-95°C.

      4.11  Graduated  cylinder or equivalent.

5.0   REAGENTS

      5.1   Reagent  Water:    Reagent water  will  be  interference free.   All
references  to  water in  this  method refer  to reagent water  unless  otherwise
specified.

      5.2   Aqua regia:  Prepare immediately before use by carefully adding three
volumes of concentrated  HC1 to one volume of concentrated HN03.

      5.3   Sulfuric acid, 0.5 N:   Dilute 14.0 ml  of concentrated  sulfuric acid
to 1 liter.

      5.4   Stannous sulfate:   Add 25 g  stannous sulfate to 250  ml  of 0.5 N
sulfuric acid.   This mixture is  a  suspension  and  should be stirred continuously
during use. A  10% solution of stannous chloride  can be substituted for stannous
sulfate.

      5.5   Sodium chloride-hydroxylamine sulfate solution:   Dissolve 12 g of
sodium chloride and  12  g  of hydroxylamine sulfate in reagent water  and dilute to
100 ml.   Hydroxylamine  hydrochloride  may be used  in place  of  hydroxylamine
sulfate.

      5.6   Potassium permanganate,  mercury-free,  5% solution (w/v):  Dissolve
5 g of potassium permanganate in 100 ml of reagent water.

      5.7   Mercury stock solution:  Dissolve 0.1354 g of mercuric chloride in
75 ml of reagent water.   Add  10 ml of  concentrated nitric acid and adjust the
volume to 100.0 ml (1.0 ml = 1.0 mg Hg).

      5.8   Mercury working standard:   Make  successive  dilutions  of  the stock
mercury solution to obtain a working standard  containing 0.1 ug/mL.   This working
standard and the  dilution of the  stock mercury solutions should be prepared fresh
daily.  Acidity  of  the working  standard should  be maintained at  0.15% nitric
acid.    This acid should be added  to the flask,   as needed,  before adding the
aliquot.

6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   All  samples  must  have been  collected using  a sampling  plan  that
addresses the  considerations discussed in Chapter Nine of this manual.

      6.2   All sample  containers  must  be prewashed with detergents, acids, and
reagent water.   Plastic and glass  containers are both suitable.

      6.3   Non-aqueous  samples   shall  be  refrigerated, when  possible,  and
analyzed as soon as  possible."
                                  7471A  - 3                       Revision 1
                                                                  September 1994

-------
7.0   PROCEDURE

      7.1   Sample preparation:  Weigh  triplicate  0.2-g portions of untreated
sample and place in the bottom of a BOD bottle.  Add 5 ml of reagent water and
5 ml of aqua regia.  Heat 2 min in a water bath at 95°C.  Cool; then add 50 ml
reagent water and 15 ml potassium permanganate solution to each sample bottle.
Mix thoroughly and place in the water bath for 30 min at 95°C.   Cool and add 6
mL of sodium chloride-hydroxylamine sulfate to  reduce the excess permanganate.

            CAUTION: Do this addition  under a hood,  as C12 could be evolved.
            Add 55 ml of reagent water.  Treating each bottle individually, add
            5 mL of  stannous sulfate and  immediately  attach  the bottle to the
            aeration apparatus.  Continue as described under step 7.4.

      7.2   An alternate digestion  procedure  employing an autoclave may also be
used.  In this method, 5 ml of concentrated  H2S04 and 2 ml of concentrated HN03
are added to the 0.2 g of sample.  Add 5 ml of saturated KMn04 solution and cover
the bottle with a piece of aluminum foil.  The samples are autoclaved  at  121°C
and 15 Ib for 15 min.  Cool,  dilute  to a volume of 100 ml  with reagent water, and
add 6 ml of sodium chloride-hydroxylamine  sulfate solution to reduce the excess
permanganate.  Purge the dead air space and continue as described under step 7.4.
Refer to the caution statement in section 7.1 for the proper protocol  in reducing
the excess permanganate solution and adding stannous sulfate.

      7.3   Standard preparation:  Transfer 0.0-, 0.5-, 1.0-, 2.0-, 5.0-, and 10-
mL aliquots of the mercury  working  standard,  containing 0-1.0 ug of mercury, to
a series of 300-mL BOD bottles  or equivalent.   Add enough reagent water to each
bottle to make a total volume of 10 ml.   Add 5 mL of aqua regia and heat 2 min
in a water bath at 95°C.  Allow the sample to cool; add 50 mL reagent water and
15 mL of KMn04 solution  to each bottle   and  return   to the  water bath for 30
min.  Cool  and  add 6 mL of sodium  chloride-hydroxylamine  sulfate solution to
reduce the  excess  permanganate.   Add 50  mL  of reagent water.   Treating each
bottle individually, add 5 mL of stannous sulfate solution,  immediately attach
the bottle to the aeration apparatus,  and continue as described in
Step 7.4.

      7.4   Analysis:  At  this  point,  the sample  is  allowed  to stand quietly
without  manual  agitation.   The  circulating  pump,  which has  previously been
adjusted to a rate of 1  L/min,  is allowed  to run continuously.  The absorbance,
as exhibited either on the  spectrophotometer  or the recorder, will increase and
reach  maximum within 30  sec.    As  soon  as  the  recorder   pen   levels  off
(approximately 1  min), open the bypass valve and continue the aeration until the
absorbance  returns  to its  minimum  value.   Close the  bypass  valve,  remove the
fritted tubing from the BOD bottle, and continue the aeration.

      7.5   Construct a calibration curve by plotting the absorbances of stan-
dards versus micrograms of mercury.  Determine the peak height of the unknown
from the chart and read the  mercury  value  from  the standard curve.  Duplicates,
spiked samples, and check standards should be routinely analyzed.

      7.6   Calculate  metal  concentrations:    (1)  by  the  method  of  standard
additions,  (2) from a calibration curve, or (3) directly from the instrument's
concentration read-out.  All dilution or  concentration factors must be taken into

                                   7471A - 4                       Revision 1
                                                                  September 1994

-------
account.   Concentrations  reported  for multiphased  or wet  samples  must  be
appropriately qualified  (e.g., 5 ug/g dry weight).

8.0   QUALITY CONTROL

      8.1  Refer to section 8.0 of Method 7000.

9.0   METHOD PERFORMANCE

      9.1   Precision and accuracy  data  are available in Method 245.5 of Methods
for Chemical Analysis of Water and Wastes.

      9.2   The data shown  in  Table  1 were  obtained from records of state and
contractor laboratories.   The  data are intended to  show the  precision of the
combined sample preparation and analysis method.

10.0  REFERENCES

1.    Methods  for  Chemical  Analysis of  Water  and  Wastes,   EPA-600/4-82-055,
December 1982, Method 245.5.

2.    Gaskill, A.,  Compilation and Evaluation of RCRA Method  Performance Data,
Work Assignment No. 2,  EPA Contract No.  68-01-7075, September 1986.
                                  7471A - 5                       Revision 1
                                                                  September 1994

-------
                      TABLE  1. METHOD  PERFORMANCE  DATA
Sample                          Preparation            Laboratory
Matrix                           Method               Replicates
Emission control dust          Not  known              12,  12 ug/g

Wastewater treatment sludge    Not  known           0.4,  0.28 ug/g
                                  7471A - 6                       Revision 1
                                                                  September 1994

-------
                              METHOD 7471A
MERCURY IN  SOLID OR SEMISOLID WASTE (MANUAL COLD-VAPOR  TECHNIQUE)
                          (   -"   )
           Sample Preparation
Standard Preparation
/ Typ
f Dige
>v Met!
1
r
9 of \TV2P<>
lod? >/
Type
f 1
7.1 Weigh triplicate
samples, and reagent
water and
aqua regia.
1
r
7.1 Heat, cool,
add reagent water
and KMn04.
1
>


>
r
7.3 Transfer aliquot*
of Hg working
standards to
bottle*.

,
r
7.3 Add reagent
water to volume,
and aqua regia,
heat and cool.
r
7.2 Add
KMn04, cover,
heat and cool,
dilute with
reagent water.
L
7.1 Heat, cool,
add sodium
chlonde-
hydroxylamme
sulfate.


I
7.2 Add sodium
chlonde-
hydroxylamine
sulfate, purge
dead air space.
i
7.1 Add reagent
water, stannou*
sulfate, attach
to aeration
apparatus.
w


\r
7.4 Analyze
sample.

1
7.5 Construct
calibration
curve; determine
peak height and
Hg value.

1
7.5 Routinely
analyze duplicates,
spiked samples.




J
r
7.3 Add reagent
water and KMnO4
solution, heat
and cool.
4
7.3 Add sodium
chloride-
hydroxylamme
sulfate and
reagent water.
i
7.C
stannoi
appi
'
Add
s sulfate,
o aeration
iratus.

                              7.6 Calculate
                                 metal
                              concentrations.
                             (    Stop   j

                               7471A  -  7
                          Revision  1
                          September 1994

-------
                                 METHOD 7480
              MOLYBDENUM (ATOMIC ABSORPTION,  DIRECT  ASPIRATION)

1.0  SCOPE AND APPLICATION
     1.1  See Section 1.0 of Method 7000.

2.0  SUMMARY OF METHOD
     2.1  See Section 2.0 of Method 7000.
3.0  INTERFERENCES
     3.1  See Section 3.0 of Method 7000 if interferences are suspected.
     3.2  Interferences in an air/acetylene flame from Ca, Sr, $04, and Fe are
severe.  These interferences are  greatly  reduced  in the nitrous oxide flame
and by addition of 1,000 mg/L aluminum to samples and standards.
4.0  APPARATUS AND MATERIALS
     4.1  For basic apparatus,  see Section 4.0 of Method 7000.
     4.2  Instrument parameters  (general):
          4.2.1  Molybdenum hollow cathode lamp.
          4.2.2  Wavelength:  313.3 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxidant:  Nitrous oxide.
          4.2.5  Type of  flame:  Fuel  rich.
          4.2.6  Background correction:  Required.
5.0  REAGENTS
     5.1  See Section 5.0 of Method 7000.
     5.2  Preparation of  standards;
          5.2.1  Stock  solution:    Dissolve  1.840  g  of ammonium molybdate,
     (NH4)sMo7024*4H20  (analytical reagent grade),  in Type II water and dilute
     to 1 liter; 1  mL  =  1  mg  Mo   (1,000  mg/L).  Alternatively, procure a
     certified standard from a  supplier and verify by comparison with a second
     standard.
                                   7480 -  1
                                                         Revision
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution   to   be   used  as
     calibration standards at the time of analysis.   The  calibration standards
     should  be  prepared  using  the  same  type   of  add  and   at   the  same
     concentration  as  will   result  1n  the  sample  to be   analyzed after
     processing.  The samples  and  standards  should  also  contain 1,000  mg/L
     aluminum (see Paragraph  5.2.3).

          5.2.3  Aluminum nitrate solution:   Dissolve  139 g aluminum  nitrate,
     A1(N03)3*9H20, in 150 ml  of  Type  II  water;  heat to  effect solution.
     Allow to cool and make up  to  200  mL.    To each 100  ml of  standard and
     sample alike, add 2 ml of the aluminum nitrate solution.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3, Sample Handling  and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method  7000.
9.0  METHOD  PERFORMANCE

     9.1  The performance characteristics for an aqueous sample free of inter-
ferences are:

     Optimum concentration  range:   1-40 mg/L with a wavelength of 313.3 nm.
     Sensitivity:   0.4 mg/L.
     Detection  limit:  0.1  mg/L.

     9.2  In a   single   laboratory,  analysis  of  a mixed industrial-domestic
waste  effluent,  digested with Method 3010,  at concentrations of 0.3, 1.5, and
7.5 mg/L gave standard deviations   of  +0.007, +0.02, and +0.07, respectively.
Recoveries at these levels  were  100%,  96%,  and 95%, respectively.

     9.3  For   concentrations  of   molybdenum  below  0.2  mg/L,  the  furnace
technique  (Method  7481)  is  recommended.


10.0 REFERENCES

1.   Methods for  Chemical  Analysis   of  Water  and Wastes, EPA-600/4-82-055,
December 1982,  Method 246.1.

                                   7480 - 2
                                                         Revision      0
                                                          Date  September  1986

-------
                  METHOD 7ABO

MOLYBDENUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
               C
                 5.O
                     Prepare
                    standards
                 7.1
                       For
                      sample
                preparation see
                    chapter 3.
                   section 3.2
                 7.2
                 Analyze using
                  Method 7OOO.
                  Section 7.Z
               (     Stop      J
            7480 -  3
                                      Revision       0
                                      Date   September 1986

-------
                                 METHOD 7481

              MOLYBDENUM (ATOMIC ABSORPTION, FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Molybdenum is  prone  to  carbide  formation.    Use a pyrolytically
coated graphite tube.

     3.3  Memory effects are  possible,  and  cleaning  of  the furnace may be
required after analysis of more concentrated samples or standards.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Drying time and temp:  30  sec at 125*C.
          4.2.2  Ashing time and temp:  30  sec at 1400*C.
          4.2.3  Atomizing time and temp:   5 sec at 2800*C.
          4.2.4  Purge gas:  Argon (nitrogen should not be used).
          4.2.5  Wavelength:  313.3 nm.
          4.2.6  Background correction:  Required.
          4.2.7  Other operating parameters should be  set as specified by the
                 particular instrument manufacturer.
          4.2.8  Pyrolytically coated graphite tube.
          NOTE:  The above concentration values  and instrument conditions are
                 for a Perkin-Elmer  HGA-2100,  based  on  the  use of a 20-uL
                 injection,  continuous-flow   purge   gas,  and  nonpyrolytic
                 graphite.    Smaller  sizes   of  furnace  devices  or  those
                 employing faster rates of  atomization  can be operated using
                 lower atomization temperatures for  shorter time periods than
                 the above-recommended settings.
                                  7481 - 1
                                                         Revision
                                                         Date  September 1986

-------
5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:    Dissolve  1.840  g  of ammonium molybdate,
     (NH4)6M07024'4H20  (analytical  reagent  grade),  1n  Type  II  water and
     dilute to 1 liter; 1 mL = 1 mg Mo (1,000 mg/L).  Alternatively, procure a
     certified standard from a supplier and verify by comparison with a second
     standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  acid  and  at  the same
     concentrations as in the sample after processing (0.5% v/v
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000,  Paragraph 7.3,  Furnace Procedure.  The calculation
is given in Method  7000,  Paragraph 7.4.


8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method  7000.


9.0  METHOD  PERFORMANCE

     9.1   Precision and accuracy data  are  not available at  this  time.

     9.2   The performance characteristics  for an  aqueous sample  free of Inter-
 ferences  are:

     Optimum concentration range:  3-60 ug/L.
     Detection  limit:   1 ug/L.


 10.0 REFERENCES

 1.   Methods for  Chemical  Analysis  of  Water  and Wastes,  EPA-600/4-82-055,
 December 1982,  Method 246.2.


                                   7481 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
9861
              U01SLA9H
                                        £ -
                              (       103S     j

9'f. u
'QOOL P<
Bui sn
azAiBu^


3T?aas
[natBD
naas
3U53H
1 Z' L

Z * C uo T ^ 3 as
'aa*jui «ja^dBu^
aas uoi 3BJ6da.jd
aiduiBS joj


I • L

pjepueis
uoi3Bjq^iBD


Q'\

OnOINHDBJL
                                  'NOIldHOSBV DIH01V)  HnN3O9A~IOH
                                 1SVL OOH13H

-------
                                 METHOD 7520

                NICKEL (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction is required.

     3.3  High concentrations  of  iron,  cobalt,  or  chromium may interfere,
requiring either matrix matching or use of a nitrous-oxide/acetylene flame.

     3.4  A nonresonance line of Ni  at 232.14 nm causes nonlinear calibration
curves at moderate to high nickel concentrations, requiring sample dilution or
use of the 352.4-nm line.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Nickel hollow cathode lamp.
          4.2.2  Wavelength:  232.0 nm (primary); 352.4 nm (alternate).
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:   Dissolve  1.000  g nickel metal  (analytical
     reagent grade)  or  4.953   g  nickel  nitrate,  Ni(N03)2*6H20  (analytical
     reagent grade), in 10 mL HNOs and   dilute   to 1 liter with Type II water.


                                   7520 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
     Alternatively,  procure a certified standard  from a  supplier  and  verify by
     comparison with a second standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution   to  be  used as
     calibration standards at the time of analysis.  The calibration  standards
     should  be  prepared  using  the  same  type  of acid and   at   the same
     concentration  as  will  result  in  the  sample to  be   analyzed  after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000,  Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics for an aqueous sample free of inter-
ferences are:

     Optimum concentration  range:  0.3-5 mg/L with a wavelength of 232.0 nm.
     Sensitivity:   0.15 mg/L.
     Detection  limit:  0.04 mg/L.

     9.2   In a   single  laboratory,  analysis  of  a mixed industrial-domestic
waste effluent,  digested with Method 3010,  at concentrations of 0.2,  1.0, and
5.0 mg/L gave  standard deviations  of   +0.011, +0.02, and +0.04, respectively.
Recoveries at  these levels  were  100%,  97%,  and 93%, respectively.

     9.3   The  data  shown  in Table   1   were  obtained  from records of state and
contractor laboratories.   The data are  intended   to  show the precision of the
combined sample preparation and  analysis method.
                                   7520 - 2
                                                          Revision
                                                          Date   September 1986

-------
10.0 REFERENCES

1.   Methods for  Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December 1982, Method 249.1

2.   Gaskm, A., Compilation and Evaluation  of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075, September 1986.
                                   7520 - 3
                                                          Revision
                                                          Date  September 1986

-------
                      TABLE 1. METHOD PERFORMANCE DATA
       Sample                      Preparation                Laboratory
       Matrix                        Method                   Replicates


Wastewater treatment sludge           3050                 13,000,  10,400 ug/g
                                   7520 - 4
                                                          Revision      0
                                                          Date  September 1986

-------
               METHOD  7520

NICKEL  (ATOMIC ABSORPTION. DIRECT ASPIRATION)
                 Start
             5.0
                 Prepare
                standards
7. 1
P^epar
ct
sec
For
samp le
atlon see
apter 3.
tion 3.2
7.2

Ana] yze
Method
Section
us Ing
7000,
7.2
                 Stop
          7520 - 5
                                  Revision       o
                                  Date  September 1986

-------
                                 METHOD 7550

                OSMIUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
1.0  SCOPE AND APPLICATION

     1.1  Method  7550  is  an   atomic   absorption  procedure  approved  for
determining  the  concentration  of   osmium  1n  wastes,  mobility  procedure
extracts, soils, and  ground  water.    All  samples  must  be subjected to an
appropriate dissolution step prior to analysis.


2.0  SUMMARY OF METHOD

     2.1  Prior to analysis  by  Method  7550,  samples  must  be prepared for
direct aspiration.  The method  of  sample  preparation will vary according to
the sample matrix.    Aqueous  samples  are  subjected  to  the acid digestion
procedure discussed in this  method.    Sludge  samples are prepared using the
procedure described in Method 3050.   For samples containing oils, greases, or
waxes, the procedure described in Method  3040  may be applicable.  Due to the
very volatile nature of some  osmium  compounds, the applicability of a method
to a sample must be verified by  means of spiked samples or standard reference
materials, or both.

     2.2  Following the appropriate dissolution  of  the sample, a representa-
tive aliquot is aspirated into a nitrous oxide/acetylene flame.  The resulting
absorption of hollow  cathode  radiation  will  be  proportional to the osmium
concentration.  Background correction must be employed for all analyses.

     2.3  The typical detection limit  for  this  method  is 0.3 mg/L; typical
sensitivity 1s 1 mg/L.


3.0  INTERFERENCES

     3.1  Background correction is required because nonspecific absorption and
light scattering can be significant at the analytical wavelength.

     3.2  Due to the volatility of osmium,  standards  must be made on a daily
basis, and the applicability of sample-preparation techniques must be verified
for the  sample matrices of Interest.

     3.3  Samples and standards should  be monitored for viscosity differences
that may  alter the aspiration rate.

     3.4  Osmium and its  compounds  are   extremely  toxic; therefore, extreme
care must be taken to ensure  that  samples and standards are handled properly
and that  all exhaust gases are properly vented.
                                   7550 - 1
                                                          Revision      0
                                                          Date  September  1986

-------
4.0  APPARATUS AND MATERIALS

     4.1  Atomic  absorption  spectrophotometer;    Single-  or  dual-channel,
single- ordouble-beamInstrumentwithagrating monochromator, photomul-
tlpller detector, adjustable slits, and provisions for background correction.

     4.2  Osmium hollow cathode lamp.

     4.3  Strip-chart recorder  (optional).


5.0  REAGENTS

     5.1  ASTM Type  II water  (ASTM  D1193):    Water  should be monitored for
impurities.

     5.2  Concentrated  nitric  acid   (HN03):    Add  should  be   analyzed to
determine levels of  impurities.IT" a  method  blank using the acide 1s 
-------
     6.4  Nonaqueous  samples  shall  be   refrigerated,   when  possible,   and
analyzed as soon as possible.


7.0  PROCEDURE

     7.1  Sample preparation;  Aqueous samples should be prepared according to
the procedure describedTrT Paragraph  7.2.    Sludge-type  samples should be
prepared according to Method 3050;  samples containing oils, greases, or waxes
may be prepared according  to  Method  3040.    The  applicability of a sample
preparation technique to a new  matrix  type must be demonstrated by analyzing
spiked samples, relevant standard reference materials, or both.

     7.2  Sample preparation of aqueous samples;

          7.2.1  Transfer a representative  100-mL  aliquot  of the well-mixed
     sample to a Griffin beaker and add 1 ml of concentrated HN03.

          7.2.2  Place the beaker on a steam bath or hot plate and warm for 15
     mln.  Cool the beaker and,  1f  necessary, filter or centrifuge to remove
     Insoluble material.

          7.2.3  Add 1 ml of concentrated H2S04  and adjust the volume back to
     100 ml.  The  sample 1s now ready for analysis.

     7.3  The 290.0-nm  wavelength  line  and  background  correction shall be
employed.

     7.4  A fuel-rich nitrous oxide/acetylene  flame shall be  used.

     7.5  Follow   the  manufacturer's  operating   Instructions  for  all other
Instrument parameters.

     7.6  Either   (1)  run   a  series  of   osmium  standards  and  construct  a
calibration curve  by plotting the   concentrations  of the standards against the
absorbances,  or   (2)  for   the  method  of standard  additions,  plot  added
concentration versus  absorbance.    For  Instruments  that   read  directly in
concentration,  set the curve corrector to read out the proper concentration.

     7.7  Analyze  all EP extracts,  all samples analyzed as part of a deli sting
petition, and all  samples that suffer  from matrix interferences by the method
of standard additions.

     7.8  Duplicates, spiked samples, and   check standards should be routinely
analyzed.

     7.9  Calculate metal  concentrations:     (1)  by  the  method of standard
additions,  (2)  from a calibration curve, or (3) directly from the Instrument's
concentration read-out.  All dilution  or   concentration factors must be taken
into account.
                                   7550 - 3
                                                         Revision      0
                                                         Date  September 1986

-------
8.0  QUALITY CONTROL

     8.1  All quality control data should be maintained and available for easy
reference or Inspection.

     8.2  Calibration curves must be  composed  of  a  minimum  of a blank and
three standards.   A  calibration  curve  should  be  made  for  every hour of
continuous sample analysis.

     8.3  Dilute samples  1f  they  are  more  concentrated  than  the highest
standard or 1f they fall on the plateau of a calibration curve.

     8.4  Employ a minimum  of  one  blank  per  sample  batch to determine If
contamination or any memory effects are occurring.

     8.5  Verify calibration  with  an  Independently  prepared check standard
every 15 samples.

     8.6  Run one spike duplicate sample  for  every  10 samples.  A duplicate
sample  1s  a  sample  brought  through  the  entire  sample  preparation  and
analytical process.

     8.7  The method of  standard  additions  (see  Method  7000, Section 8.7)
shall be used for the analysis  of  all EP extracts, on all analyses submitted
as part of a del 1 sting  petition,  and  whenever  a new sample matrix 1s being
analyzed.


9.0  METHOD  PERFORMANCE

     9.1   Precision  and accuracy data are not available at  this  time.


 10.0 REFERENCES

 1.   Methods for   Chemical   Analysis  of  Water  and Wastes,  EPA-600/4-82-055,
 December 1982,  Method 252.1.
                                   7550 - 4
                                                          Revision
                                                          Date  September 1986

-------
                                         METHOD 7SSO

                    OSMIUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION METHOD)
                         C
        Samples containing
          oils,  greases.
            or wanes
                 Sludge-type
	   ^ * wwv «: w 71-
                   samples
  ror sampje     '
 preparation
    aliquot of
     sample to
   beaker:  add
    cone. HN03
                                                    Use Method 3050
Use Method 3040
                            Warm Beaker;
                          cool and filter
                            If necessary
                              Add cone.
                           H,SO«:  adjust
                               volume
                                                    7.3-5
                                                                                   Adjust
                                                                                 Instrument
                                                                                 parameters
                                                                               7.6
    Plot
callbrat ion
   curve
                                                                               7.7
                                                          Analyze
                                                      by method of
                                                         standard
                                                        additions
                                                                               7.B
                                                         Routinely
                                                          analyze
                                                       duplicates.
                                                    spiked samples
                                                        and check
                                                        standards
                                                                               7.9
                                                                              Calculate natal
                                                                              concentrations
                                                                             f      Stop      J
                                     7550 -  5
                                                                Revision        0
                                                                Date  September 1986

-------
                                 METHOD 7610

              POTASSIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  In air/acetylene or other  high-temperature flames (>2800*C),  potas-
sium can experience  partial  ionization,  which indirectly affects absorption
sensitivity.  The presence of other alkali salts in the sample can reduce this
ionization and thereby enhance analytical results.  The ionization-suppressive
effect of sodium is small if the ratio  of  Na to K is under 10.  Any enhance-
ment due to sodium can be stabilized  by adding excess sodium (1,000 ug/mL) to
both sample and standard solutions.    If more stringent control of ionization
is required, the addition  of  cesium  should  be  considered.  Reagent blanks
should be analyzed to correct for potassium impurities in the buffer stock.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Potassium hollow cathode lamp.
          4.2.2  Wavelength:  766.5 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxidant:  Air.
          4.2.5  Type of flame:  Slightly oxidizing  (fuel lean).
          4.2.6  Background correction:  Not required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.
                                  7610 - 1
                                                         Revision
                                                         Date  September 1986

-------
     5.2  Preparation of standards;

          5.2.1  Stock solution:   Dissolve  1.907   g  of  potassium chloride,  KC1
     (analytical  reagent grade),  dried at 110*C  1n Type II water and  dilute to
     1 liter with Type II water.    Alternatively,  procure a  certified standard
     from a supplier and verify by comparison  with a second  standard.

          5.2.2  Prepare  dilutions  of  the  stock   solution   to  be  used as
     calibration standards at the time of analysis.   The calibration  standards
     should  be  prepared  using  the  same  type  of  acid  and  at   the same
     concentration  as  will   result  In  the   sample  to  be   analyzed  after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3, Sample Handling  and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation:  The  procedures  for   preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.
9.0  METHOD PERFORMANCE

     9.1  The performance characteristics for an aqueous sample free of Inter-
ferences are:

     Optimum concentration range:  0.1-2 mg/L with a wavelength of 766.5 nm.
     Sensitivity:  0.04 mg/L.
     Detection  limit:  0.01 mg/L.

     9.2  In a   single  laboratory,  analysis  of  a mixed Industrial-domestic
waste  effluent,  digested with Method   3010,  at  concentrations of 1.6 and 6.3
mg/L gave standard deviations of  +0.2 and +0.5, respectively.  Recoveries at
these  levels were 103% and 102%,  respectively.


10.0 REFERENCES

1.   Methods for Chemical  Analysis   of  Water  and Wastes,  EPA-600/4-82-055,
December  1982,  Method 258.1.
                                   7610 - 2
                                                          Revision
                                                          Date   September 1986

-------
                  METHOD 761O

POTASSIUM (ATOMIC ABSORPTION.  DIRECT  ASPIRATION)
                  5.O
                      Prepare
                     standards
7. 1
prepar
cl-
sec
For
sample
atton see
apter 3.
tlon 3.2
                  7.2
                  Analyze using
                   Method 7000.
                   Section  7.2
                f      Stop       J
               7610 -  3
                                         Revision       0
                                         Date  September  1986

-------
                                 METHOD 7740

               SELENIUM (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  Method  7740  Is  an   atomic   absorption  procedure  approved  for
determining  the  concentration  of  selenium  in  wastes,  mobility-procedure
extracts, soils, and  ground  water.    All   samples  must  be subjected to an
appropriate dissolution step prior to analysis.


2.0  SUMMARY OF METHOD

     2.1  Prior to analysis by Method 7740,   samples must be prepared in order
to convert organic forms of  selenium  to inorganic forms, to minimize organic
interferences, and to convert samples to suitable solutions for analysis.  The
sample-preparation procedure varies, depending on  the sample matrix.  Aqueous
samples are  subjected  to  the  acid-digestion  procedure  described  in this
method.  Sludge samples are  prepared  using the procedure described in Method
3050.

     2.2  Following the appropriate dissolution  of  the sample, a representa-
tive aliquot 1s placed manually  or  by  means  of an automatic sampler into a
graphite tube furnace.    The  sample  aliquot  is  then  slowly evaporated to
dryness, charred  (ashed),  and  atomized.    The  absorption of lamp radiation
during atomization will be proportional to the selenium concentration.

     2.3  The typical detection limit for this method is 2 ug/L.


3.0  INTERFERENCES

     3.1  Elemental   selenium  and  many   of   Its  compounds  are  volatile;
therefore,  samples   may  be   subject  to  losses  of  selenium  during sample
preparation.  Spike samples  and  relevant standard reference materials should
be processed to determine if the chosen dissolution method 1s appropriate.

     3.2  Likewise,   caution   must  be   employed   during  the  selection  of
temperature and times for the  dry  and  char   (ash) cycles.  A nickel nitrate
solution must   be added  to   all   dlgestates  prior  to  analysis to minimize
volatilization  losses during drying and ashing.

     3.3  In addition to the   normal  interferences  experienced during graphite
furnace  analysis,  selenium   analysis  can  suffer from severe  nonspecific
absorption  and  light  scattering   caused   by   matrix  components   during
atomization.  Selenium analysis  is  particularly susceptible  to these problems
because  of  its  low analytical  wavelength (196.0 nm).  Simultaneous  background
correction  is required to avoid  erroneously   high  results.  High iron  levels
can  give   overcorrection  with   deuterium  background.    Zeeman   background
correction  can  be useful in this situation.
                                   7740 - 1
                                                          Revision      0
                                                          Date   September  1986

-------
     3.4  If the analyte 1s  not  completely  volatilized and removed from the
furnace during atomlzatlon, memory effects  will   occur.   If this situation 1s
detected, the tube should be cleaned by operating the furnace at full power at
regular intervals In the analytical scheme.

     3.5  Selenium analysis suffers  interference  from  chlorides (>800 mg/L)
and sulfate (>200 mg/L).  The  addition  of nickel nitrate such that the final
concentration is 1% nickel will lessen this interference.


4.0  APPARATUS AND MATERIALS

     4.1  250-mL Griffin beaker.

     4.2  10-mL volumetric flasks.

     4.3  Atomic  absorption  spectrophptometer;    Single-  or  dual-channel,
single- or double-beam  instrument  with  a grating monochromator, photomulti-
plier detector,  adjustable  slits,  a  wavelength  range  of  190-800 nm, and
provisions for  simultaneous  background  correction  and  interfacing  with a
strip-chart recorder.

     4.4  Selenium hollow  cathode  lamp, or electrode! ess discharge lamp (EDL);
EDLs provide better sensitivity for the analysis of Se.

     4.5  Graphite furnace;  Any graphite  furnace device with the appropriate
temperature and timing  controls.

     4.6  Strip-chart   recorder:    A  recorder   is  strongly  recommended for
furnace work so that there will be a permanent record and so that any problems
with the  analysis,  such  as  drift,  Incomplete  atomization,  losses during
charring, changes in sensitivity,  etc., can easily be recognized.

     4.7  Pi pets;  Microliter with disposable tips.  Sizes can range from
5  to 1,000  uL,  as required.


5.0  REAGENTS

     5.1  ASTM Type  II  water  (ASTM   D1193):     Water  should  be monitored for
impurities.
      5.2  Concentrated  nitric  acid  (HNOs);     Acid   should   be   analyzed  to
 determine levels of impurities.   Tf~a method blank made with the acid is  
-------
     5.4  Selenium standard stock  solution  (1,000  mg/L):    Either procure a
certified aqueous standard fromasupplier  and  verify by comparison with a
second standard, or dissolve 0.3453  g  of  selenlous add (actual  assay 94.6%
H2Se03, analyticaTreagent grade) or equivalent in Type II water and dilute to
200 mL.

     5.5  Nickel nitrate solution  (5%):    Dissolve  24.780  g of ACS reagent
grade Ni(N03)2*6H20 or equivalent in Type II water and dilute to 100 ml.

     5.6  Nickel nitrate solution (1%):  Dilute 20 ml of the 5% nickel nitrate
to 100 ml with Type II water.

     5.7  Selenium working standards;  Prepare dilutions of the stock solution
to be used as calibration  standards  at  the  time of the analysis.  Withdraw
appropriate aliquots  of the stock solution,  add  1 ml of  concentrated HN03,
2 ml  of 30% H202,  and 2 ml  of the 5%  nickel nitrate  solution.   Dilute to
100 ml with Type II water.

     5.8  Air;  Cleaned and  dried  through  a  suitable filter to remove oil,
water, and other foreign substances.    The  source  may  be a compressor or a
cylinder of industrial-grade compressed air.

     5.9  Hydrogen;  Suitable for instrumental analysis.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  All samples must  have  been  collected  using  a sampling plan that
addresses the considerations discussed in Chapter Nine of this manual.

     6.2  All sample containers must be  prewashed with detergents, acids, and
Type II water.  Plastic and glass containers are both suitable.

     6.3  Special  containers  (e.g.,  containers  used  for  volatile organic
analysis) may have to be used  1f  very  volatile selenium compounds are to be
analyzed.

     6.4  Aqueous samples must be acidified to a pH of <2 with nitric acid.

     6.5  Nonaqueous  samples  shall  be   refrigerated,  when  possible,  and
analyzed as soon as possible.


7.0  PROCEDURE

     7.1  Sample preparation;    Aqueous  samples  should  be  prepared 1n the
manner described inSteps7.1.1  to  7.1.3.    Sludge-type samples should be
prepared according to Method 3050.   The applicability of a sample-preparation
technique to a  new  matrix  type  must  be  demonstrated  by analyzing spiked
samples and/or  relevant standard reference materials.
                                  7740 - 3
                                                         Revision
                                                         Date  September 1986

-------
          7.1.1   Transfer 100  mL   of   well-mixed   sample  to  a 250-mL Griffin
     beaker;  add  2 ml of 30%   ^2 and  sufficient concentrated HN03 to  result
     in an acid concentration  of  1%  (v/v).   Heat  for  1  hr at  95*C  or until  the
     volume is slightly less than 50  mL.

          7.1.2   Cool and bring back  to  50 ml with Type II water.

          7.1.3   Pi pet 5 ml of this   digested solution  into a 10-mL volumetric
     flask, add  1 ml of the  1%  nickel   nitrate  solution, and  dilute  to 10 ml
     with Type II water.    The  sample   is  now   ready for injection  into the
     furnace.

     7.2  The 196.0-nm wavelength line and a background correction system must
be employed.  Follow the  manufacturer's suggestions  for all  other spectropho-
tometer parameters.

     7.3  Furnace parameters suggested by  the manufacturer should be employed
as  guidelines.    Because   temperature-sensing   mechanisms   and  temperature
controllers can vary between  instruments  or  with  time,  the  validity of the
furnace parameters must be  periodically  confirmed by systematically altering
the furnace parameters while analyzing a  standard.  In this  manner,  losses of
analyte due to overly high  temperature  settings or losses in  sensitivity due
to less than  optimum  settings  can  be  minimized.    Similar verification of
furnace parameters may be required for complex sample matrices.

     7.4   Inject  a measured uL-al1quot of sample  into the furnace and atomize.
If the concentration found is  greater  than  the highest standard, the sample
should be diluted in the same acid matrix and reanalyzed.  The use of multiple
injections can improve accuracy  and  help detect furnace pipetting errors.

     7.5  Analyze all  EP extracts, all samples analyzed as part of a deli sting
petition,  and all  samples that suffer from matrix interferences by the method
of standard  additions.

     7.6   Run a  check  standard after approximately every 10 sample Injections.
Standards  are run in part to monitor the  life and performance of the graphite
tube.   Lack  of reproducibility   or   significant   change  in the signal  for the
standard  indicates that  the tube should be replaced.

     7.7   Duplicates,  spiked  samples, and  check   standards should be analyzed
every  20  samples.

     7.8   Calculate  metal  concentrations:     (1)  by  the  method of standard
additions,  (2) from  a  calibration curve,  or  (3) directly from the  instrument's
concentration read-out.  All  dilution or concentration factors must be  taken
into account.
                                   7740 - 4
                                                          Revision
                                                          Date  September 1986

-------
8.0  QUALITY CONTROL

     8.1  All quality control data should be maintained and available for easy
reference or inspection.

     8.2  Calibration curves must be  composed  of  a  minimum  of a blank and
three standards.   A  calibration  curve  should  be  made  for  every hour of
continuous sample analysis.

     8.3  Dilute samples  if  they  are  more  concentrated  than  the highest
standard or  if they fall on the plateau of a calibration curve.

     8.4  Employ a minimum  of  one  blank  per  sample  batch to determine if
contamination or any memory effects are occurring.

     8.5  Verify calibration  with  an  independently  prepared check standard
every 15 samples.

     8.6  Run one spike duplicate sample  for  every  10 samples.  A duplicate
sample  is   a  sample  brought  through  the  entire  sample  preparation  and
analytical process.

     8.7  The method of  standard  additions  (see  Method  7000, Section 8.7)
shall be used for the analysis  of  all EP extracts, on all analyses submitted
as  part of a deli sting  petition,  and  whenever  a new sample matrix is being
analyzed.


9.0 METHOD  PERFORMANCE

     9.1  Precision and accuracy data are available in Method  270.2  of Methods
for Chemical Analysis of Water  and Wastes.

     9.2  The data  shown in  Table  1  were  obtained from  records of state and
contractor laboratories.   The data are  intended  to show  the  precision of the
combined sample  preparation  and analysis method.


10.0 REFERENCES

1.   Methods for Chemical   Analysis  of  Water   and Wastes, EPA-600/4-82-055,
December 1982, Method 270.2.

2.   Gaskill, A., Compilation and Evaluation  of  RCRA Method Performance Data,
Work Assignment  No. 2,  EPA Contract No. 68-01-7075, September  1986.
                                   7740 - 5
                                                          Revision       0
                                                          Date  September  1986

-------
                      TABLE 1. METHOD PERFORMANCE DATA
       Sample                      Preparation                Laboratory
       Matrix                        Method                   Replicates


Emission control dust                 3050                    14, 11 ug/g
                                   7740 - 6
                                                          Revision
                                                          Date  September 1986

-------
                 METHOD 7740

SELENIUM (ATOMIC ABSORPTION.  FURNACE  METHOD)
                Type  of  sample
                  for sample
                 preparat ion
               7.1 . 11  Transfer
                    1  portion
                 of  sample  to
               beaker;  add 30X
                HzOi  and  cone.
                   HNO,
                         heat
               7. 1.2
                Cool:  bring to
                    volume
               7.1.3
                      Plpet
                     digested
                     solution
               into  flask; add
                nickel nitrate
               solution:dilute
Sludge-type
  samples
         7. l
         Prepare  sample
          according  to
          Method  3050
                7740 - 7
                                           Revision       0
                                           Date   September 1986

-------
                                        METHOD 7740

                       SELENIUM (ATOMIC  ABSORPTION  FURNACE METHOD)
                                        (Cont inued)
7.2
Set Instrument
  parameters
7.3
                                                                             7.5
                                                         Analyze
                                                        method of
                                                        standard
                                                        addition
  Periodically
check validity
   of furnace
   parameters
7.4
                                                   7.6  I

                                                         Run
                                                   check standard
                                                  after 10 sample
                                                     Injections
Inject sample
into furnace:
   atomize
     Is
concentration
  > highest
  standard?
Dilute sample
and reanalyze
                                                                             7.7
                                                        Routinely
                                                         analyze
                                                      duplicates.
                                                  spiked samples.
                                                      and check
                                                      standards
                                                                             7.8
Calculate metal
 concentrations
                                                                           (     Stop       J
                                     7740 -  8
                                                                Revision       0
                                                                Date  September  1986

-------
                                 METHOD 7741A

                 SELENIUM (ATOMIC ABSORPTION,  GASEOUS HYDRIDE)
1.0   SCOPE AND APPLICATION

      1.1   Method 7741 is an atomic absorption procedure that is approved for
determining the concentration of selenium in wastes, mobility-procedure extracts,
soils, and ground water, provided that the sample matrix does not contain high
concentrations of chromium,  copper, mercury, silver, cobalt,  or molybdenum.  All
samples must be subjected to an  appropriate dissolution  step  prior to analysis.
Spiked  samples  and  relevant  standard  reference materials  are employed  to
determine applicability of  the  method  to a  given  waste.  If interferences are
present the analyst should consider using Method  7740.

2.0   SUMMARY OF METHOD

      2.1   Samples are prepared according to the nitric/sulfuric  acid digestion
procedure described  in this method.   Next,  the  selenium in  the digestate is
reduced to Se(IV) with  tin chloride.  The Se(IV) is then  converted to  a volatile
hydride  with  hydrogen produced  from  a zinc/HCl  or  sodium  borohydrate/HCl
reaction.

      2.2   The volatile hydride is swept into an argon-hydrogen flame located
in the optical  path  of an  atomic  absorption  spectrophotometer;  the resulting
absorbance is proportional to the selenium concentration.

      2.3   The typical detection limit  for this method is 0.002 mg/L.

3.0   INTERFERENCES

      3.1   High concentrations  of chromium, cobalt, copper, mercury, molybdenum,
nickel, and silver can cause analytical  interferences.

      3.2   Traces of  nitric acid left following the sample work-up can result
in analytical  interferences.  Nitric acid must be distilled  off  the sample by
heating the sample until fumes of S03 are observed.

      3.3   Elemental selenium and many of its compounds are volatile; therefore,
certain samples may be  subject to losses  of selenium during sample preparation.

4.0   APPARATUS AND MATERIALS

      4.1   100-mL beaker.

      4.2   Electric  hot  plate  or  equivalent -  Adjustable  and  capable  of
maintaining a temperature of 90-95°C.

      4.3   A commercially available zinc slurry hydride generator  or a generator
constructed from the following material  (see Figure 1):
                                   7741A  -  1                       Revision 1
                                                                  September 1994

-------
            4.3.1    Medicine  dropper:   Fitted into a  size  "0"  rubber stopper
      capable of delivering 1.5 ml.

            4.3.2    Reaction  flask:  50-mL,  pear-shaped,  with  two 14/20 necks
      (Scientific Glass, JM-5835).

            4.3.3  Gas  inlet-outlet tube:  Constructed  from a micro cold-finger
      condenser  (JM-3325) by  cutting the  portion  below the  14/20 ground-glass
      joint.

            4.3.4   Magnetic stirrer:   To homogenize the zinc slurry.

            4.3.5   Polyethylene drying  tube:   10-cm, filled with glass wool to
      prevent particulate matter from entering the burner.

            4.3.6   Flow meter:  Capable of measuring 1 liter/min.

      4.4   Atomic absorption  spectrophotometer: Singleordual channel, single-
or double-beam instrument with  a grating monochromator, photomultipl ier detector,
adjustable  slits,   a wavelength  range  of  190-800 nm,  and  provisions  for
interfacing with a  strip-chart recorder  and simultaneous background correction.

      4.5   Burner:  Recommended by the particular instrument manufacturer for
the argon-hydrogen flame.

      4.6   Selenium hollow cathode lamp or electrodeless discharge lamp.

      4.7   Strip-chart recorder (optional).

5.0   REAGENTS

      5.1   Reagent water:  Water should be monitored for impurities.  Reagent
water will be interference free.   All  references to water  will refer to reagent
water.

      5.2   Concentrated  nitric  acid:   Acid  should be analyzed  to  determine
levels of impurities.  If a method blank  made with the acid is  
-------
      5.7   Stannous  chloride solution:   Dissolve 100  g  SnCl2 in  100  ml of
concentrated HC1.

      5.8   Selenium  standard stock  solution:    1,000  mg/L  solution  may be
purchased, or prepared as follows:  Dissolve 0.3453 g of selenious acid  (assay
94.6% of H2Se03) in reagent water.  Add to a 200-mL volumetric flask and bring
to volume (1 ml =  1 mg Se).

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must have  been  collected using a  sampling  plan  that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All sample containers must be prewashed with  detergents,  acids, and
reagent water.  Plastic and glass containers are both suitable.

      6.3   Special  containers   (e.g.,  containers used  for  volatile  organic
analysis) may  have to be used  if very volatile selenium compounds  are to be
analyzed.

      6.4   Aqueous samples must be acidified to a pH of <2 with nitric acid.

      6.5   Nonaqueous samples shall be refrigerated, when possible,  and analyzed
as soon as possible.

7.0   PROCEDURE

      7.1   Sample preparation:

            7.1.1   To a 50-mL  aliquot  of digested sample  (or,  in  the case of
      extracts, a 50-mL sample)   add  10 mL of   concentrated HN03   and 12 mL of
      18 N H2S04.   Evaporate the sample on a hot  plate  until  white S03 fumes are
      observed  (a  volume of  about 20  mL).  Do not let  it char.   If it chars,
      stop the digestion, cool,  and add additional  HN03.  Maintain  an excess of
      HN03 (evidence of brown  fumes) and  do not let the solution darken because
      selenium may be reduced and  lost.   When the sample remains  colorless or
      straw yellow during evolution of S03 fumes,  the  digestion is  complete.

            Caution: Venting reaction vessels should be done with
            caution and only under a fume hood or well ventilated
            area.

            7.1.2   Cool the  sample,  add about  25 mL  reagent  water,  and again
      evaporate to S03 fumes just to expel oxides of nitrogen.  Cool.  Add 40 mL
      concentrated HC1 and bring to a volume of  100 mL with  reagent water.

      7.2   Prepare working standards from the  standard  stock solutions.   The
following procedures provide standards in the optimum range.

            7.2.1   To  prepare  a working stock solution, pipet  1  mL standard
      stock solution (see Paragraph 5.8) into a 1-liter volumetric flask.  Bring
      to volume with  reagent  water  containing 1.5 mL  concentrated HN03/liter.
      The  concentration  of this solution is 1  mg Se/L (1 mL  = 1  ug Se).

                                   7741A  -  3                       Revision 1
                                                                  September  1994

-------
            7.2.2   Prepare six working standards by transferring 0, 0.5, 1.0,
      1.5, 2.0, and 2.5 ml of the working stock solution (see Paragraph 7.2.1)
      into  100-mL  volumetric  flasks.    Bring to  volume with  diluent.   The
      concentrations of these working standards are  0,  5, 10, 15, 20, and 25 ug
      Se/L.

      7.3   Standard additions:

            7.3.1     Take  the 15-,  20-,   and 25-ug  standards and  transfer
      quantitatively 25 ml from each into separate 50-mL volumetric flasks.  Add
      10 ml of the  prepared sample to each.  Bring to volume with reagent water
      containing 1.5 ml HN03/liter.

            7.3.2   Add 10  ml  of  prepared  sample  to a 50-mL volumetric flask.
      Bring to volume with reagent water containing  1.5 ml HN03/liter.  This is
      the blank.

      7.4   Follow  the manufacturer's  instructions for  operating   an  argon-
hydrogen flame.  The  argon-hydrogen flame is  colorless; therefore,  it  may be
useful to  aspirate  a  low  concentration of sodium to  ensure  that ignition has
occurred.

      7.5   The 196.0-nm wavelength shall  be used  for the analysis of selenium.

      7.6   Transfer a 25-mL portion of the digested sample or standard to the
reaction vessel. Add 0.5 ml SnCl2  solution.  Allow at least  10 min for the metal
to be reduced to its lowest oxidation state.   Attach the reaction vessel to the
special  gas  inlet-outlet  glassware.   Fill  the medicine dropper with  1.50 ml
sodium borohydrate  or  zinc slurry  that  has  been  kept  in  suspension  with the
magnetic stirrer.  Firmly insert the stopper containing the medicine dropper into
the side neck  of the reaction  vessel.   Squeeze the  bulb to introduce the zinc
slurry or  sodium borohydrate  into  the  sample  or  standard solution.   The metal
hydride will produce a peak almost immediately.  When the recorder pen returns
partway to the base line,  remove the reaction vessel.

8.0   QUALITY CONTROL

      8rl  Refer to section 8.0 of Method 7000.

9.0   METHOD PERFORMANCE

      9.1   Precision and  accuracy data are available in Method 270.3 of Methods
for Chemical Analysis of Water and Wastes.

10.0  REFERENCES

1.    Methods  for  Chemical  Analysis  of  Water and  Wastes,  EPA-600/4-82-055,
December 1982, Method 270.3.
                                   7741A -  4                       Revision 1
                                                                  September 1994

-------
                          METHOD  7741A
     SELENIUM  (ATOMIC  ABSORPTION,  GASEOUS HYDRIDE)
                         C   Start    J
          Standard Preparation
          Sample Preparation
  7.2.1  Pip.t
     • toek
 •olution into
 flack;  bring
   to volume
7.2.2 Prepare 6
  £• working
•tandarda from
•took;  bring to
    volvun
7.3.1 Tranafer
  3 standard
 portion*,add
•ample,bring to
    volume
                            7.1.1 Add
                          concentrated
                         H.SO.  and UNO.
                          to cample and
                            avaporata
  7.1.1 Stop
digeation,cool,
   add HMO.
                           7.1.2 Cool
                           •ample,add
                         reagent water,
                         evaporate,cool
   7.3.2 To
 prepare blank
add aampla to a
flaak and bring
   to volume
  7.4 Fallow
 instruction*
 for operating
irgon-hydrogen
     flame
 7.5 Uae  196.0
 nm wavelength
                            7.1.2 Add
                          concentrated
                          HC1 and bring
                            to volume
 7.6 Tranafer
digevted aamplfl
  to reaction
  veaael,add
     SnCl,
                          7.6 Allow to
                          •tand,attach
                           vaaael to
                          glauware ,add
                           2n alurry
                         7.6 Record Se
                         concentration
C
                             Stop
                            7741A  - 5
                                             Revision  1
                                             Septenter 1994

-------
                                  METHOD 7742

              SELENIUM (ATOMIC ABSORPTION,  BOROHYDRIDE REDUCTION)
1.0  SCOPE AND APPLICATION

      1.1  Method 7742 is an atomic absorption  procedure  for  determining 3
to 750 fjg/l concentrations of selenium in wastes, mobility procedure extracts,
soils,  and  ground water.   Method 7742  is  approved for  sample  matrices  that
contain  a  total  of  up  to 1000  mg/L  concentrations of  cobalt,  copper,  iron,
mercury, and  nickel. A  solid  sample  can contain up  to  10% by  weight  of the
interferents  before  exceeding 1000 mg/L in a  digested  sample.   All  samples
including aqueous matrices must be subjected to  an appropriate dissolution step
prior to analysis.  Spiked samples and relevant  standard reference materials are
employed to determine the applicability of the method to a given waste.

2.0  SUMMARY OF METHOD

      2.1  Samples are prepared according to the nitric acid digestion procedure
described   in   Method   3010   for  aqueous  and   extract   samples   and   the
nitric/peroxide/hydrochloric acid digestion procedure described  in Method 3050
(furnace AA  option)   for  sediments,  soils,   and  sludges.   Excess  peroxide is
removed  by  evaporating   samples  to near-dryness at  the  end of  the digestion
followed by dilution to  volume and degassing the samples upon addition of urea.
The selenium  is  converted  to the +4 oxidation  state  during  digestion  in  HC1.
After a 1:10 dilution, selenium is then converted to  its volatile hydride using
hydrogen  produced from  the  reaction of  the  acidified  sample  with  sodium
borohydride in a continuous-flow hydride generator.

      2.2  The volatile  hydrides  are   swept into,  and decompose in,  a heated
quartz  absorption  cell  located  in the optical  path of an  atomic absorption
spectrophotometer.    The  resulting  absorption  of  the  lamp  radiation  is
proportional to the  selenium concentration.

      2.3  The typical detection limit for this method is 3//g/L.

3.0  INTERFERENCES

      3.1   Very  high (>1000  mg/L)  concentrations  of cobalt,  copper,  iron,
mercury, and,  nickel  can cause analytical  interferences through precipitation as
reduced metals and associated blockage of transfer lines and fittings.

      3.2  Traces of peroxides left following the sample work-up can result in
analytical  interferences.  Peroxides must be removed  by evaporating each sample
to  near-dryness   followed  by  reacting  each sample  with  urea and  allowing
sufficient time for degassing before analysis (see Sections 7.1 and 7.2).

      3.3   Even  after acid digestion, flame gases  and organic  compounds  may
remain  in  the  sample.   Flame  gases and  organic  compounds  can  absorb  at  the
analytical  wavelengths and background correction should be used.

                                    7742-1                        Revision 0
                                                                  September  1994

-------
4.0  APPARATUS AND MATERIALS

      4.1  Electric  hot  plate:   Large enough to hold at  least  several  100 mL
Pyrex digestion beakers.

      4.2   A  continuous-flow hydride  generator:    A commercially  available
continuous-flow  sodium  borohydride/HCl  hydride  generator  or  a  generator
constructed similarly to that shown in Figure 1 (P. S. Analytical or equivalent).

            4.2.1  Peristaltic Pump:  A four-channel, variable-speed peristaltic
      pump to permit regulation of liquid-stream flow rates (Ismatec Reglo-100
      or equivalent).  Pump  speed and tubing diameters should  be  adjusted to
      provide  the  following  flow rates:    sample/blank  flow  = 4.2  mL/min;
      borohydride flow = 2.1 mL/min.

            4.2.2  Sampling Valve (optional):     A sampling valve (found in the
      P. S.  Analytical  Hydride  Generation  System  or equivalent)  that  allows
      switching between samples and blanks (rinse solution) without introduction
      of air into the system will provide more signal  stability.

            4.2.3  Transfer Tubing and Connectors: Transfer tubing (1 mm I.D.),
      mixing T's, and connectors are made of fluorocarbon (PFA or TFM) and are
      of  compatible   sizes  to  form  tight,  leak-proof  connections  (Latchat,
      Technicon, etc. flow injection apparatus accessories or equivalent).

            4.2.4  Mixing  Coil:  A 20-turn coil made by wrapping transfer tubing
      around a 1-cm diameter by 5-cm long plastic or glass rod (see Figure 1).

            4.2.5  Mixing Coil  Heater,  if appropriate:   A  250-mL  Erlenmeyer
      flask containing 100 mL  of water heated to boiling  on  a  dedicated one-
      beaker hotplate (Corning PC-35 or equivalent).   The mixing coil  in 4.2.4
      is immersed in  the  boiling  water to  speed kinetics of the hydride forming
      reactions  and   increase  solubility   of  interfering   reduced   metal
      precipitates.

            4.2.6  Gas-Liquid Separator:  A glass apparatus for collecting and
      separating liquid  and  gaseous products  (P.  S.  Analytical  accessory or
      equivalent) which allows the liquid fraction to drain to waste and gaseous
      products above the liquid to be swept by a regulated carrier gas (argon)
      out of the cell for  analysis.   To  avoid undue carrier  gas dilution, the
      gas volume above the liquid should  not  exceed 20 mL.  See Figure 1 for an
      acceptable separator shape.

            4.2.7  Condenser:   Moisture  picked up by the  carrier gas  must be
      removed before encountering the  hot  absorbance cell.   The moist carrier
      gas with the  hydrides is dried by passing the gasses  through a small  (< 25
      mL) volume condenser coil  (Ace  Glass Model  6020-02 or equivalent) that is
      cooled to 5°C by a water chiller  (Neslab RTE-110 or equivalent).  Cool tap-
      water in place of a chiller is acceptable.
                                    7742-2                        Revision 0
                                                                  September 1994

-------
            4.2.8  Flow Meter/Regulator:   A meter  capable of regulating up to 1
      L/min of argon carrier gas is recommended.

      4.3  Absorbance Cell:  A 17-cm or longer quartz  tube T-cell  (windowless is
strongly suggested)  is recommended, as shown  in Figure 1  (Varian Model  VGA-76
accessory or equivalent).  The  cell  is  held in place  by a holder that positions
the cell  about 1  cm over a  conventional  AA air-acetylene  burner head.   In
operation, the cell  is heated to around 900°C.

      4.4   Atomic  absorption  spectrophotometer:    Single-  or  dual-  channel,
single- or double-beam instrument having a grating monochromator, photomultiplier
detector, adjustable slits, a wavelength  range of  190 to 800 nm, and provisions
for interfacing with  an appropriate recording device.

      4.5  Burner:   As recommended by the  particular instrument manufacturer for
an air-acetylene flame.   An appropriate mounting bracket attached to the burner
that suspends the quartz  absorbance cell  between  1  and 2 cm above  the burner slot
is required.

      4.6  Selenium hollow cathode lamp or selenium electrodeless discharge lamp
and  power  supply.   Super-charged  hollow-cathode  lamps  or   EDL  lamps  are
recommended for maximum  sensitivity.

      4.7     Strip-chart  recorder   (optional):       Connect   to   output  of
spectrophotometer.

5.0  REAGENTS

      5.1  Reagent  water :   Water must be monitored  for  impurities.   Refer to
Chapter 1 for definition of Reagent water.

      5.2  Concentrated nitric acid  (HN03):  Acid  must be analyzed to determine
levels of impurities.  If a method blank is 
-------
                               QUARTZ CELL

                               A  A   OURNER
                                                                     TO
                                                                   CHILLER
 '•01SCONNECTE
  OUR INO S*X8n
 .  ntlOLVStS
                                                              OASXLIOUIO
                                                              SEPARATOR
                                                               __—# DRAIN
                       20 TURN COIL
                         (TEFLON)
                         HOTPLATE
                         VALVE .
                        (•LANK)
Figure 1.  Continuous-flow sodium borohydride/hydride generator apparatus setup
and an AAS sample introduction  system
                                     7742-4
Revision 0
September 1994

-------
      5.7  4% Sodium Borohydride (NaBHJ:  A 4 % sodium borohydride solution (20
g reagent-grade NaBH4 plus 2 g sodium hydroxide dissolved in 500 mL of reagent
water) must be prepared for conversion of the selenium to its hydride.

      5.8  Selenium solutions:

            5.8.1   Selenium  standard stock  solution  (1,000  mg/L):   Either
      procure  certified  aqueous  standards  from  a  supplier   and  verify  by
      comparison with a second standard,  or dissolve  0.3453 g of selenious acid
      (assay 96.6% of H2Se03)  in 200  ml  of reagent water (1 ml  =  1 mg Se).

            5.8.2   Selenium  working stock solution:   Pi pet  1 ml selenium
      standard stock solution  into all volumetric  flask  and  bring to volume
      with  reagent water containing  1.5  mL concentrated  HN03/liter.    The
      concentration of this solution is  1 mg Se/L (1  ml = 1 fjq  Se).

6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples must  have been  collected  using a sampling  plan that
addresses the considerations discussed in Chapter Nine of this  manual.

      6.2  All sample containers must be prewashed with detergents,  acids, and
reagent water.  Plastic and glass containers are both suitable.

      6.3   Special  containers  (e.g.,  containers  used  for volatile  organic
analysis) may have to be used  if very volatile selenium compounds are suspected
to be present in the samples.

      6.4  Aqueous samples must be acidified to a pH  of <2  with nitric acid.

      6.5  Nonaqueous samples shall be refrigerated, when possible, and analyzed
as soon as possible.

7.0  PROCEDURE

      7.1  Place a 100-mL  portion of  an aqueous sample or extract or 1.000 g of
a dried solid sample in a 250-mL digestion beaker.   Digest  aqueous samples and
extracts according to Method 3010. Digest solid samples according  to Method 3050
(furnace AA option) with the following modifications:  add 5 mL of concentrated
hydrochloric acid  just prior   to  the final volume reduction  stage  to  aid  in
conversion of selenium  to  its plus four state;  the final volume reduction should
be to less than  5  mL  but  not  to dryness  to  adequately  remove  excess hydrogen
peroxide (see note).  After dilution to volume, further dilution with diluent may
be necessary if the analyte is known to exceed 750 /jg/L or  if interferents are
expected to exceed  a total of 1000  mg/L in the digestate.

            Note:  For  solid digestions,  the volume reduction stage is critical
            to  obtain  accurate  data.    Close monitoring  of  each   sample  is
            necessary when this critical  stage in  the digestion is reached.
                                    7742-5                        Revision 0
                                                                  September 1994

-------
      7.2  Prepare samples for hydride analysis by adding 1.00 g  urea, and 20 mL
concentrated HC1 to a 5.00 ml aliquot of digested sample in a 50-mL volumetric
flask.  Heat in a water bath  to  dissolve  salts and reduce selenium (at least 30
minutes  is suggested).    Bring flask  to  volume  with  reagent water  before
analyzing.    A  ten-fold  dilution   correction   must  be  made  in  the  final
concentration calculations.

      7.3  Prepare working standards  from the standard stock selenium solution.
Transfer  0, 0.5,  1.0,  1.5, 2.0, and  2.5 ml of  standard to  100-mL  volumetric
flasks and bring to volume with diluent.  These concentrations will be 0, 5, 10,
15, 20, and 25/yg Se/L.

      7.4   If  EP extracts (Method  1310)  are being  analyzed  for selenium,  the
method of standard additions must be used.  Spike appropriate  amounts of working
standard  selenium  solution to  three  25 ml  aliquots of  each  unknown.   Spiking
volumes should be  kept less than 0.250  ml to  avoid  excessive spiking dilution
errors.

      7.5   Set up  instrumentation and  hydride generation apparatus  and fill
reagent containers.  The  sample  and blank flows should  be set  around 4.2 mL/min,
and the  borohydride flow  around 2.1  mL/min.    The  argon carrier gas  flow is
adjusted to about 200 mL/min.  For the AA,  use the 196.0-nm wavelength and 2.0-nm
slit width (or manufacturer's recommended  slit-width)  with background correction.
Begin all flows and allow the  instrument  to  warm-up  according to the instrument
manufacturer's instructions.

      7.6  Place sample feed line into a prepared sample  solution and start pump
to begin  hydride generation.    Wait  for  a maximum  steady-state signal  on  the
strip-chart recorder.  Switch to blank sample and watch for  signal to decline to
baseline  before  switching  to the next  sample  and beginning the  next analysis.
Run standards  first (low to  high),  then  unknowns.   Include  appropriate QA/QC
solutions, as required.   Prepare calibration curves  and convert absorbances to
concentration.  See following analytical  flowchart.

      CAUTION: The hydride of selenium is very toxic.  Precautions must be taken
      to  avoid inhaling the gas.

      7.7  If the method of standard additions was employed,  plot the measured
concentration  of the  spiked  samples  and unspiked  sample versus  the  spiked
concentrations.  The spiked concentration axis intercept will be the method of
standard  additions  concentration.   If the plot  does  not result in  a straight
line,   a   nonlinear  interference is  present.   This  problem  can  sometimes be
overcome  by dilution or  addition of  other reagents  if there  is  some knowledge
about the waste.  If the method of standard additions was not  required, then the
concentration is determined from a standard calibration curve.
8.0  QUALITY CONTROL

      8.1  Refer to Section 8.0 of Method 7000.
                                    7742-6                        Revision 0
                                                                  September 1994

-------
9.0  METHOD PERFORMANCE

      9.1  The relative standard deviation obtained by a single laboratory for
7 replicates of a contaminated soil was 18% for selenium at 8.2 ug/L in solution.
The average percent recovery of  the analysis of an 2 jjg/L  spike on  ten different
samples is 100.5% for selenium.

10.0  REFERENCES

1.    Methods  for  Chemical  Analysis  of  Water  and Wastes,   EPA-600/4-82-055,
      December 1982, Method 206.3.

2.    "Evaluation of Hydride Atomic Absorption  Methods  for Antimony,  Arsenic,
      Selenium, and Tin",  an EMSL-LV internal report under Contract 68-03-3249,
      Job Order  70.16,  prepared for  T.  A. Hinners by D.  E.  Dobb, and J.  D.
      Lindner of Lockheed Engineering and Sciences Co.,  and L. V. Beach of the
      Varian Corporation.
                                    7742-7                        Revision 0
                                                                  September 1994

-------
                                 METHOD  7742
SELENIUM  (ATOMIC  ABSORPTION,  BOROHYDRIDE  REDUCTION)
    7.1 Use Method
   3050 (furnace AA
   option) to digest
     1.0 g sample.
       7.1 - 7.4
      Digest with
       H203as
      described in
     Method 3050.
        7.5 Add
      concentrated
          HCI.
       7.6 Do final
         volume
      reduction and
       dilution, as
       described.
       7.1 Further
       dilute with
         diluent.
                                7.1 Use
                              Method 3010
                              to digest 100
                               ml sample.
  7.2 Add urea
and cone. HCI to
 aliquot; heat in
   H20 bath;
bring to volume.
  7.3 Prepare
    working
 standards from
 standard stock
  Se solution.
                                7.4 Use the
                                method of
                                 standard
                                additions on
                               extracts, only.
                                    7.4 Spike 3
                                    aliquote with
                                      working
                                    standard Se
                                     solution.
                              7.5 -7.6 Analyze
                                the sample
                               using hydride
                                generation
                                apparatus.
7.5 - 7.6 Analyze
  the sample
  using hydride
  generation
  apparatus.
7.7 Determine
     Se
concentrations
  from linear
    plot.
                                   7.7 Determine
                                   Se cone, from
                                     standard
                                    calibration
                                      curve.
                                       Stop
                                        7742-8
                                               Revision  0
                                               September 1994

-------
                                 METHOD 7760A

                 SILVER  (ATOMIC ABSORPTION. DIRECT ASPIRATION)
1.0  SCOPE AND APPLICATION

      1.1 Method 7760 is an atomic absorption procedure approved for determining
the concentration of silver  (CAS  Registry Number 7440-22-4) in wastes, mobility
procedure extracts,  soils, and ground water. All samples must  be subjected to an
appropriate dissolution step prior to analysis.


2.0  SUMMARY OF METHOD

      2.1   Prior to analysis by Method 7760, samples must be prepared  for direct
aspiration.  The method of sample preparation will  vary according to the sample
matrix. Aqueous samples are  subjected  to the acid-digestion procedure described
in this method.

      2.2   Following the appropriate dissolution of the sample, a representative
aliquot is aspirated  into an air/acetylene flame. The resulting absorption of
hollow  cathode radiation will  be proportional to  the  silver  concentration.
Background correction must be employed for all analyses.

      2.3   The typical detection limit for this method is 0.01 mg/L; typical
sensitivity is 0.06 mg/L.


3.0  INTERFERENCES

      3.1  Background correction  is required because nonspecific absorption and
light scattering may occur at the analytical wavelength.

      3.2  Silver nitrate solutions are light-sensitive and have  the tendency to
plate out on  container  walls. Thus silver  standards  should be stored in brown
bottles.

      3.3  Silver chloride  is insoluble; therefore, hydrochloric acid should be
avoided unless the silver is already in solution as  a chloride complex.

      3.4  Samples and standards should be monitored for viscosity differences
that may alter the aspiration rate.


4.0  APPARATUS AND MATERIALS

      4.1 Atomic absorption spectrophotometer: Single-or dual-channel, single-
or double-beam instrument with a grating monochromator,  photomultiplier detector,
adjustable slits, and provisions for background correction.

      4.2  Silver hollow cathode lamp.


                                   7760A  -  1                       Revision 1
                                                                  July 1992

-------
      4.3  Strip-chart recorder (optional).

      4.4  Graduated cylinder or equivalent.

      4.5  Hot  plate  or equivalent - adjustable and capable  of maintaining a
temperature of 90-95°C.

      4.6   Ribbed watchglasses or equivalent.


5.0  REAGENTS

      5.1  Reagent grade chemicals shall  be used  in all tests. Unless otherwise
indicated, it is intended that all  reagents shall conform to the specifications
of the Committee on Analytical Reagents of  the American Chemical Society, where
such specifications are  available. Other grades may be used, provided it is first
ascertained that the reagent  is of sufficiently  high  purity to permit its use
without lessening the accuracy of the determination.

      5.2  Reagent Water.  Reagent water is interference free.   All references to
water in the method refer to  reagent water unless otherwise specified.

      5.3  Nitric Acid  (concentrated), HN03.

      5.4  Ammonium Hydroxide (concentrated), NH4OH.

      5.5  Silver Stock Standard Solution  (1,000 mg/L), AgN03. Dissolve 0.7874
g anhydrous silver nitrate in water.  Add  5  ml HNO, and  bring to volume  in a 500-
mL volumetric flask (1 ml  = 1  mg Ag). Alternatively,  procure a  certified aqueous
standard from a supplier and  verify by comparison with a second standard.

      5.6  Silver working  standards  - These standards  should be prepared from
silver  stock solution  to be used as  calibration  standards  at the  time  of
analysis. These standards  should  be  prepared  with nitric  acid and at the same
concentrations as the analytical solution.

      5.7  Iodine solution (IN). Dissolve  20 g potassium iodide  (KI), in 50 ml
of water. Add 12.7 g iodine (I2) and dilute to 100 ml. Store in  a brown bottle.

      5.8   Cyanogen  iodide   solution. Add 4.0  ml  ammonium  hydroxide,  6.5 g
potassium cyanide (KCN), and  5.0 ml of iodine solution to 50 ml of water. Mix and
dilute to 100 ml with water.  Do not keep longer than 2 weeks.

      CAUTION:    This  reagent cannot be mixed with any acid  solutions because
                  toxic hydrogen cyanide will be produced.

      5.9  Air.

      5.10  Acetylene.
                                   7760A - 2                      Revision  1
                                                                  July  1992

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples must  have been collected  using  a sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2  All sample containers must be prewashed with detergents, acids, and
reagent water. Plastic and glass containers are both suitable.

      6.3  Aqueous samples must be acidified to a pH < 2 with nitric acid.

      6.4  When  possible,  standards and samples should be  stored  in  the dark and
in brown bottles.

      6.5  Nonaqueous samples shall be refrigerated, when  possible,  and analyzed
as soon as possible.


7.0  PROCEDURE

      7.1  Sample preparation - Aqueous  samples should  be prepared  according to
Steps 7.2 and 7.3. The applicability  of  a  sample preparation technique to a new
matrix type  must  be demonstrated by analyzing spiked  samples and/or relevant
standard reference materials.

      7.2  Preparation of aqueous samples

            7.2.1  Transfer a representative aliquot of the well-mixed sample to
      a beaker and add 3 mL  of  concentrated HNO,.  Cover the beaker with a ribbed
      watch glass. Place the beaker on a hot plate and cautiously  evaporate to
      near dryness, making certain that the sample does not boil.  DO NOT BAKE.
      Cool the beaker and add  another 3-mL portion of  concentrated HN03. Cover
      the beaker  with a watch  glass  and  return to the hot  plate.   Increase the
      temperature of the hot plate so that a gentle reflux action  occurs.

      NOTE:  If  the  sample  contains  thiosulfates, this  step  may result  in
            splatter  of  sample  out  of  the  beaker  as the  sample approaches
            dryness. This has  been reported to occur with certain  photographic
            types of samples.

            7.2.2  Continue heating, adding additional  acid, as necessary, until
      the digestion is complete (generally indicated when  the digestate is light
      in  color  or does  not change  in  appearance with  continued  refluxing).
      Again, evaporate to  near  dryness and cool the beaker. Add a small quantity
      of HN03 so that the final dilution contains 0.5% (v/v)  HN03  and warm the
      beaker to dissolve  any precipitate  or residue resulting  from  evaporation.

            7.2.3  Wash down the  beaker  walls  and watch  glass with water and,
      when necessary, filter the sample  to remove  silicates and other insoluble
      material  that  could  clog  the nebulizer.  Adjust  the  volume  to  some
      predetermined value  based on the expected metal concentrations. The sample
      is now ready for analysis.


                                   7760A  - 3                        Revision 1
                                                                   July 1992

-------
      7.3    If  plating  out of  AgCl  is  suspected,  the  precipitate  can  be
redissolved by adding cyanogen  iodide to the sample. This can  be done only after
digestion and after neutralization of the sample to  a pH > 7  to prevent formation
of toxic cyanide under acid conditions. In  this case  do not  adjust the sample
volume to the predetermined value until  the  sample  has been neutralized to pH >
7 and cyanogen iodide has been  added.  If cyanogen iodide  addition to the sample
is necessary, then  the standards must  be  treated  in  the same manner.  Cyanogen
iodide must not be  added to the acidified silver standards. New standards must
be made, as directed  in  Steps  5.5  and  5.6,  except  that  the acid addition step
must be omitted.  For example, to obtain a 100 mg/L working standard, transfer 10
ml of stock solution to a small beaker.  Add  water to make about 70 ml.  Make the
solution  basic   (pH above  7)  with  ammonium  hydroxide. Rinse  the pH  meter
electrodes into the solution with water. Add 1 ml cyanogen iodide and allow to
stand 1 hour. Transfer quantitatively to a 100-mL volumetric  flask and bring to
volume with water.

      CAUTION;    CNI  reagent  can be  added only  after digestion  to  prevent
                  formation of toxic  cyanide  under  acidic  conditions.   CNI
                  reagent must not be added to the acidified silver standards.

      NOTE:       Once the  sample or  sample aliquot  has been treated  with the
                  CNI reagent  and diluted per  instruction,  the solution has a
                  cyanide concentration of approximately  260 mg/L. A solution of
                  that  cyanide concentration  must  be considered  a potential
                  hazardous waste  and must  be disposed of  using  an  approved
                  safety plan  in accordance with local authority requirements.
                  Until  such time  that a  detailed disposal  plan can  be fully
                  documented and approved, the use  of  the CNI reagent should be
                  avoided.

      7.4   The  328.1  nm wavelength  line  and background  correction  shall  be
employed.

      7.5   An oxidizing air-acetylene flame shall be used.

      7.6   Follow  the  manufacturer's  operating  instructions  for all  other
spectrophotometer parameters.


8.0   QUALITY CONTROL

      8.1  Refer to section 8.0 of Method 7000.
9.0   METHOD PERFORMANCE

      9.1   Precision and accuracy data are available in Method 272.1 of "Methods
for Chemical Analysis of Water and Wastes."

      9.2   The data  shown  in  Table  1  were obtained from records of  state and
contractor  laboratories.  The data are  intended  to show the  precision of the
combined sample preparation  and analysis method.


                                   7760A -  4                       Revision 1
                                                                   July  1992

-------
10.0  REFERENCES

1.    Methods for  Chemical  Analysis of  Water and Wastes;  U.S.  Environmental
Protection Agency.  Office of Research and Development. Environmental Monitoring
and Support  Laboratory.  ORD  Publication  Offices  of Center  for  Environmental
Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.

2.    Gaskill, A.,  Compilation and Evaluation of RCRA Method Performance Data,
Work Assignment No. 2, EPA Contract No. 68-01-7075, December 1987.

3.    Rohrbough, W.G.;  et  al.  Reagent  Chemicals,  American  Chemical  Society
Specifications,  7th ed.; American Chemical Society: Washington, DC, 1986.

4.    1985 Annual Book of ASTM Standards, Vol. 11.01; "Standard Specification for
Reagent Water";  ATSM: Philadelphia, PA, 1985; D1193-77.
                                   7760A  -  5                       Revision 1
                                                                  July 1992

-------
                                 TABLE  1.
                          METHOD  PERFORMANCE DATA
    Sample                        Preparation      Laboratory
    Matrix                          Method         Replicates
Wastewater treatment sludge          3050           2.3, 1.6 mg/Kg


Emission control dust                3050           1.8, 4.2 mg/Kg
                                                                                i
                                                                                 i
                                 7760A - 6                       Revision  1
                                                                 July 1992

-------
                                         METHOD  7760A
                   SILVER (ATOMIC  ABSORPTION, DIRECT  ASPIRATION)
  721 Transfer
 sample aliquot to
  beaker,add cone
 HNO,,evaporate to
near dryness,cool,
add cone HNO,,heat
 so gentle reflux
   action occurs
  7.2.2  Compl.t.
digeilion,«vaporate
 to near drjrna»» ,
cool,add cone HNOB,
 warm to di«*olv«
any pracipitat* or
      r»idu*
7.2 3 Filter sample
if necessary,adjust
 volume with water
  7.3 Neutralize
sample,add cyanogen
iodide to dissolve
precipitate,remake
standards omitting
   acid,transfer
 aliquot of stock
solution to beaker,
     add water
                         7.1  Prepare sample
                         according to Method
                               3040
7 1 Prepare sample
according  to Method
       3050
                         7.3 Adjust pH with
                            NH,OH.rinse
                           electrode into
                           solution with
                         water,add cyanogen
                           iodide,wait 1
                          hour,transfer to
                           flask,bring to
                          volume with water
                                                                             7.4-7.6 Set
                                                                             instrument
                                                                             parameters
                                                                            7.7 Construct
                                                                          calibration  curve
                                                                           7.8 Analy2e  by
                                                                          •thod of standard
                                                                             addition if
                                                                              necessary
                        7  9  Calculate metal
                          concentration
                                                                                Stop
                                            7760A  -  7
                                                                                      Revision  1
                                                                                      July 1992

-------
                                  METHOD  7761

                 SILVER  (ATOMIC  ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

      1.1   Method  7761  is  an  atomic  absorption  procedure  approved  for
determining the concentration of silver  in wastes, mobility procedure extracts,
soils, and  ground water.   All  samples  must be  subjected to  an  appropriate
dissolution procedure.


2.0  SUMMARY OF METHOD

      2.1   See Section 2.0 of Method 7000.


3.0  INTERFERENCES

      3.1   See Section 3.0 of Method 7000 if interferences are suspected.

      3.2   In addition to the normal  interferences experienced during graphite
furnace analysis, silver analysis can  suffer from severe nonspecific absorption
and  light   scattering   caused  by  matrix   components   during  atomization.
Simultaneous background correction must  be  employed  to  avoid erroneously high
results.

      3.3   If the analyte  is  not  completely  volatilized  and removed from the
furnace during  atomization,  memory effects  will occur.    If  this  situation is
detected,  the  tube  should  be   cleaned  by  operating  the  furnace  at  higher
atomization temperatures.

      3.4   Silver nitrate solutions are light sensitive and have the tendency
to plate  out on  container walls.  Thus, silver standards  should  be stored in
brown bottles.

      3.5   Silver chloride is insoluble; therefore,  hydrochloric acid should be
avoided unless the silver is already in solution as  a chloride complex.


4.0  APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument Parameters  (General):

            4.2.1  Drying Time and Temp:  30 sec at  125°C.

            4.2.2  Ashing Time and Temp:  30 sec at  400°C.

            4.2.3  Atomizing Time and Temp:   10 sec  at 2700°C.

                                   7761 - 1                        Revision 0
                                                                  July 1992

-------
            4.2.4  Purge Gas Atmosphere: Argon.

            4.2.5  Wavelength:  328.1 nm.

            4.2.6  Background Correction:  Required.

            4.2.7  Other operating parameters should be set as specified by the
      particular instrument manufacturer.

            NOTE: The above concentration values and instrument conditions are
                  for a  Perkin-Elmer HGA-2100,  based  on the  use  of a  20  uL
                  injection,  continuous  flow  purge  gas  and  non-pyrolytic
                  graphite and are to be used as guidelines only.  Smaller size
                  furnace devices or those employing faster rates of atomization
                  can be  operated  using lower  atomization  temperatures  for
                  shorter time periods than the  above recommended settings.


5.0  REAGENTS

      5.1  See Section 5.0 of Method 7000.

      5.2  Silver Stock Standard Solution (1,000 mg/L), AgN03.  Dissolve 0.7874 g
anhydrous  silver nitrate  (AgN03),  analytical  reagent grade,  water.   Add 5  ml
concentrated nitric acid (HN03) and bring to volume in a 500 ml volumetric flask
(1 mL = 1 mg Ag).  Alternatively, procure a certified standard from a supplier
and verify by comparison with a second standard.

      5.3  Silver working  standards  -  These standards should be prepared with
nitric acid such that the final acid concentration  is 0.5% (v/v) HN03.

      5.4  Ammonium hydroxide (concentrated), (NH4OH).  Base  should be analyzed
to determine  levels  of  impurities.  If  impurities  are detected,  all  analyses
should be blank-corrected.

      5.5  Iodine solution (IN).  Dissolve 20 g potassium iodide  (KI), analytical
reagent grade, in 50 ml water.  Add 12.7 g  iodine (I2), analytical reagent grade,
and dilute to 100 ml with water.  Store in a brown  bottle.

      5.6  Cyanogen  iodide  solution.   To 50 ml water add 4.0 ml concentrated
NH,OH,  6.5 g potassium cyanide  (KCN),  and 5.0 mL  of iodine  solution.   Mix and
dilute to  100 ml with water.  Do not keep longer than 2 weeks.

      CAUTION:    This  reagent  cannot  be mixed  with  any  acid  solutions since
                  highly toxic hydrogen cyanide will be produced.


6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See Chapter Three, Step 3.1.3, Sample Handling and Preservation.
                                   7761 - 2                       Revision 0
                                                                  July 1992

-------
      6.2   Standards and samples should be stored in the dark,  in brown bottles,
and refrigerated.


7.0  PROCEDURE

      7.1   Sample preparation - Aqueous samples should be prepared according to
Steps 7.2 and 7.3.  The applicability of a sample preparation technique to a new
matrix type  must  be  demonstrated by analyzing  spiked  samples and/or relevant
standard reference materials.

      7.2   Preparation of aqueous samples

            7.2.1  Transfer a representative aliquot of the well-mixed sample to
      a beaker and add 3  ml of concentrated HN03. Cover the beaker with a watch
      glass.  Place the beaker on the hot plate and cautiously  evaporate to near
      dryness, making certain that the sample does not boil.  DO NOT BAKE.  Cool
      the beaker and add another 3-mL portion of concentrated HN03.   Cover the
      beaker with  a  watch glass  and  return  to the  hot plate.   Increase  the
      temperature of the hot plate so that a gentle reflux action occurs.

            NOTE: If the sample contains thiosulfates, this step may result in
                  splatter of sample out of the  beaker  as the  sample approaches
                  dryness.  This has been reported to occur with certain types
                  of  photographic wastes.

            7.2.2  Continue heating, adding additional acid, as necessary, until
      the digestion is complete (generally indicated when the digestate is light
      in  color  or does  not  change in  appearance  with continued  refluxing).
      Again,  evaporate to near  dryness  and  cool   the beaker.   Add a  small
      quantity of HN03 so that the final dilution contains 0.5% (v/v) HN03 and
      warm  the  beaker to dissolve  any  precipitate  or  residue  resulting from
      evaporation.

            7.2.3  Wash  down  the beaker walls  and watch glass with  water and,
      when necessary, filter the  sample  to remove silicates and other insoluble
      material  that   could  clog  the nebulizer.  Adjust  the  volume to  some
      predetermined value based on the expected metal  concentrations.  The sample
      is now ready for analysis.

      7.3    If  plating  out   of  AgCl  is suspected,  the   precipitate  can  be
redissolved by adding cyanogen iodide to  the sample.  This can  be done  only after
digestion and after neutralization of the sample to a pH > 7 to  prevent formation
of toxic cyanide under acid  conditions.   In this case,  do not  adjust the sample
volume to the predetermined  value until  the sample has  been neutralized to pH >
7 and cyanogen iodide has been added.  If cyanogen  iodide addition to  the sample
is necessary, then the standards must be treated  in the same manner. Cyanogen
iodide must not be added  to  the acidified silver standards.  New standards must
be made,  as  directed in  Step 5.2,  except that  the  acid addition  step  must be
omitted.  For example, to obtain  a  100 mg/L working  standard,  transfer 10 ml of
stock solution  to  a  small beaker.  Add water to  make  about  70 ml.   Make the
solution basic  (pH above 7) with NH4OH.  Rinse the pH meter electrodes into the

                                   7761  - 3                       Revision 0
                                                                  July 1992

-------
solution with  water.   Add  1  ml  cyanogen  iodide and  allow  to stand  1  hour.
Transfer quantitatively to a  100-mL volumetric  flask  and  bring to volume with
water.

      CAUTION:    CNI  reagent can be  added only  after digestion  to  prevent
                  formation of  toxic  cyanide  under  acidic  conditions.    CNI
                  reagent must not be added to the acidified silver standards.

NOTE: Once the sample  or  sample  aliquot  has been treated  with  the CNI  reagent
      and diluted per  instruction, the solution has a cyanide concentration of
      approximately 260 mg/L.  A  solution of that cyanide concentration must be
      considered a potential  hazardous waste and must be  disposed of using an
      approved safety  plan in accordance  with local  authority requirements.
      Until such time that a detailed  disposal plan can  be fully documented and
      approved, the use of the CNI reagent should be avoided.

      7.4  The 328.1-nm wavelength line and background  correction shall be used.

      7.5   Following  the manufacturer's operating instructions for all  other
spectrophotometer parameters.

      7.6  Furnace parameters suggested by the manufacturer should be employed
as guidelines. Since temperature-sensing mechanisms and temperature controllers
can vary between  instruments or with time, the validity of the furnace parameters
must be periodically confirmed by systematically altering the furnace parameters
while analyzing a standard.  In this manner,  losses of analyte due to  higher than
necessary temperature settings or losses  in sensitivity due to less than optimum
settings can  be  minimized.  Similar  verification of  furnace parameters may be
required for complex sample matrices.

      7.7  Inject a measured uL aliquot of  sample into the furnace and atomize.
If the  concentration  found is greater than the highest  standard,  the sample
should be diluted in the  same acid matrix and reanalyzed.   The use of multiple
injections can improve accuracy and help detect furnace pipetting errors.

      7.8    Either  (1)  run  a  series  of  silver  standards  and  construct  a
calibration curve by plotting the concentrations of  the standards against the
absorbances or (2) for the method  of standard additions, plot added concentration
versus abosrbance.   For instruments that read directly in concentration, set the
curve corrector  to read out the proper concentration.

      7.9  Analyze, by the  method of standard  additions,  all  EP extracts, all
samples  analyzed as part  of a delisting petition,  and all samples that suffer
from matrix interferences.

      7.10  Calculate  metal  concentrations by (1)  the   method  of  standard
additions, or (2) from  a calibration curve,  or (3) directly  from  the instrument's
concentration readout.   All  dilution or concentration factors must  be taken into
account.  Concentrations reported for multiphased samples  must  be appropriately
qualified.
                                   7761 - 4                       Revision 0
                                                                  July 1992

-------
8.0  QUALITY CONTROL

      8.1   All quality control data should be maintained and  available for easy
reference or inspection.

      8.2   Calibration curves must be composed  of  a  minimum of a calibration
blank and three standards.  A calibration curve must be prepared each day.

      8.3   Dilute  samples  if  they  are  more concentrated  than  the  highest
standard or if they fall on the plateau of a calibration curve.

      8.4   Employ a minimum of one reagent blank per sample batch or every 20
samples to determine if contamination or any memory effects are occurring.

      8.5   Verify calibration with an  independently  prepared quality control
reference sample every 10 samples.

      8.6   Run  one spiked  replicate  sample for  every  10  samples  or  per
analytical batch, whichever is more frequent.  A replicate sample is a sample
brought through the entire sample preparation process.

      8.7   Duplicates, spiked samples,  and check standards should  be routinely
analyzed.  Refer to Chapter One for the proper protocol.

      8.8   The method of standard additions (see Method 7000, Step 8.7) shall
be used for the analysis of all EP extracts, on all  analyses submitted as part
of a delisting petition, and whenever a new sample matrix is being analyzed.


9.0  METHOD PERFORMANCE

      9.1   Precision and accuracy data are available in Method 272.2 of Methods
for Chemical Analysis of Water and Wastes.

      9.2   The  performance  characteristics  for an  aqueous  sample  free  of
interferences are:

      Optimum concentration range:  1-25 ug/L.
      Detection limit:  0.2 ug/L.


10.0  REFERENCES

1.    Methods  for  Chemical  Analysis of  Water and  Wastes;  U.S.  Environmental
Protection Agency. Office of Research and Development. Environmental Monitoring
and  Support  Laboratory. ORD  Publication  Offices of  Center  for Environmental
Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.
                                   7761 - 5                       Revision 0
                                                                  July 1992

-------
                                            METHOD  7761
                      SILVER  (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)
                                  7.1 Must
                                 demonstrate
                              applicability of
                                 preparation
                             technique to other
                               matrix types by
                              analyzing spiked
                                 samples and
                             reference material*
  7.2.1 Transfer
 •ample aliquot to
 beaker;  add cone.
HNOi;  evaporate to
near  dryness; cool;
  add  cone. HNO,;
  heat to gentle
   reflux action
      occur*
  7.2.2 Complete
    digestion;
 evaporate  to near
dryneas;  cool; add
cone.  HNOi;  warm to
   dissolve any
  precipitate or
      residue
                                      Ye*
7 2.3 Filter  temple
   if necessary;
adjust volume with
       water
   7.3  Neutralize
(ample;  add cyanogen
 iodide  to dissolve
 precipitate; remake
 »tandard* omitting
   acid;  transfer
  aliquot of stoek
 solution to beaker;
     add water
 7.3 Adjust pH with
    NH.OH; rinse
   electrodes into
  soln with water;
    add cyanogen
   iodide; wait 1
  hour; transfer to
   flask; bring to
  volume with water
     7.4-7.6 Sat
     instrument
     parameters
 7.7 Inject  sample
aliquot;  dilute  if
     necessary
   7.8 Construct
 calibration  curve
                           i
 7.9-7.10 Analyze
      sample
                                                                                   7.10 Calculate
                                                                                  •tal concentration
                                                                                       Stop
                                        7761  -  6
                           Revision 0
                           July 1992

-------
                                 METHOD 7770

                SODIUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 1f Interferences are suspected.

     3.2  lonlzation interferences can affect  analysis for sodium; therefore,
samples and standards  must  be  matrix  matched  or an ionization suppressant
employed.

     3.3  Sodium is a universal contaminant, and great care should be taken to
avoid contamination.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Sodium  hollow cathode lamp.
          4.2.2  Wavelength:   589.6 nm.
          4.2.3  Fuel:   Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Not required.


5.0  REAGENTS

     5.1  See Section 5.0 of  Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:    Dissolve  2.542  g   sodium  chloride, NaCl
     (analytical  reagent grade),  in   Type  II  water,  acidify  with   10 mL
     redistilled  HN03,   and   dilute   to   1    liter    with  Type   II  water.
     Alternatively, procure a certified  standard from a  supplier and verify by
     comparison with a second standard.


                                   7770 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution   to   be   used  as
     calibration standards at the time of analysis.   The calibration standards
     should  be  prepared  using  the  same  type  of  add and  at   the same
     concentration  as  will   result  1n  the  sample  to  be   analyzed  after
     processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The performance characteristics for an aqueous sample free of Inter-
ferences are:

     Optimum concentration range:  0.03-1 mg/L with a wavelength of 589.6 nm.
     Sensitivity:  0.015 mg/L.
     Detection  limit:  0.002 mg/L.

     9.2  In a   single  laboratory,   analysis  of  a mixed  Industrial-domestic
waste effluent,  digested with  Method  3010,  at  concentrations of 8.2 and 52
mg/L gave standard deviations of   +0.1   and +0.8, respectively.  Recoveries at
these levels were 102% and 100%,  respectively.


10.0 REFERENCES

1.   Methods for Chemical   Analysis  of  Water   and Wastes, EPA-600/4-82-055,
December  1982,  Method 273.1.
                                   7770 - 2
                                                          Revision
                                                          Date  September  1986

-------
                 METHOD 7770

SODIUM (ATOMIC  ABSORPTION. DIRECT ASPIRATION)
               5.0
                   Prepare
                  standards
               7. 1
                     For
                    sample
              preparation  see
                  chapter  3.
                 section 3.2
               7.2
               Analyze  using
                Method  7000.
                Section 7.3
             f      Stop       J
             7770 - 3
                                        Revision       0
                                        Date   September 1986

-------
                                 METHOD 7780

               STRONTIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0   SCOPE AND APPLICATION

      1.1   See Section 1.0 of Method 7000.


2.0   SUMMARY OF METHOD

      2.1   See Section 2.0 of Method 7000.


3.0   INTERFERENCES

      3.1   See Section 3.0 of Method 7000.

      3.2   Chemical interference caused by silicon, aluminum,  and phosphate are
controlled by adding lanthanum chloride. Potassium chloride is  added to suppress
the ionization of strontium. All samples  and  standards  should contain 1 ml of
lanthanum chloride/potassium chloride solution (Step 5.3) per 10 ml of solution.


4.0   APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument parameters (general):

            4.2.1   Strontium hollow cathode lamp.

            4.2.2   Wavelength:  460.7 nm.

            4.2.3   Fuel:  Acetylene.

            4.2.4   Oxidant:  Air.

            4.2.5   Type of flame:   Oxidizing (fuel lean).

            4.2.6   Background correction: not required.


5.0   REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards


                                   7780 -  1                        Revision 0
                                                                  July 1992

-------
            5.2.1  Stock solution:   (1.0 ml =  1.0 mg  Sr).  Dissolve  2.415 g of
      strontium nitrate, Sr(N03)2,  in 10 ml of concentrated HC1  and  700 ml of
      water. Dilute to  1 liter with  water. Alternatively,  procure  a certified
      standard from a  supplier and verify by comparison with a second standard.

            5.2.2   Prepare dilutions  of  the  stock  solution  to  be used  as
      calibration standards at the time of analysis.  The calibration standards
      should be prepared using the same type of  acid  as  the  samples and cover
      the range of expected concentrations in the  samples. Calibration standards
      should also contain 1 ml of lanthanum chloride/potassium chloride solution
      per 10 ml.

      5.3   Lanthanum Chloride/Potassium Chloride Solution. Dissolve 11.73 g of
lanthanum oxide, La203,  in a minimum amount of concentrated  hydrochloric acid
(approximately 50 ml). Add 1.91 g of  potassium  chloride, KC1. Allow solution to
cool to room temperature and dilute to 100 ml  with water.

      CAUTION:    REACTION IS  VIOLENT!  Add acid slowly and  in small portions to
                  control  the reaction rate upon  mixing.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See Chapter Three, Step 3.1.3, Sample Handling  and Preservation.


7.0   PROCEDURE

      7.1   Sample preparation - The procedures  for preparation of the sample are
given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.2,  Direct  Aspiration.


8.0   QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.
9.0   METHOD PERFORMANCE

      9.1   The  performance characteristics  for  an  aqueous  sample free  of
interferences are:

      Optimum concentration range:  0.3 - 5 mg/L at a wavelength of 460.7 nm.
      Sensitivity: 0.15 mg/L.
      Detection limit: 0.03 mg/L.

            9.1.1   Recoveries  of  known  amounts of  strontium  in a  series  of
      prepared standards were as given in Table 1.

                                   7780 - 2                       Revision 0
                                                                  July 1992

-------
10.0  REFERENCES

1.    Annual  Book of ASTM Standards; ASTM;   Philadelphia,  PA,  1983;  D3920.
                                   7780 -  3                       Revision 0
                                                                  July 1992

-------
                              TABLE 1.
                              RECOVERY

Amount
added,
mg/L

Amount
found ,
mg/L
Significant
(95 %
% confidence
Bias Bias level)
        1.00
        0.50
        0.10
          Reagent  Water Type II

      0.998        -0.002     -0.2
      0.503        -1-0.003     +0.6
      0.102        +0.002     +2

             Water of Choice
no
no
no
1.00
0.50
0.10
1.03
0.504
0.086
+0.03
+0.004
-0.014
+ 3
+ 0.8
-14
no
no
no
Reference:
Annual Book of ASTM Standards; ASTM:  Philadelphia, PA,
1983; D3920.
                               7780 -  4
                                            Revision  0
                                            July 1992

-------
                     METHOD  7780
STRONTIUM  (ATOMIC ABSORPTION,  DIRECT  ASPIRATION)
                        Start
                     S.0 Prepare
                      standard*
                    7.1 For lanple
                    preparation • ••
                  Chapter 3, Section
                         3.2
                   7.2 Analyze u»ing
                    Method 7000
                     Section 7.2
                        Stop
                     7780 - 5
Revision  0
July 1992

-------
                                 METHOD 7840

               THALLIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction is required.

     3.3  Hydrochloric add should not be used.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Thallium hollow cathode lamp.
          4.2.2  Wavelength:  276.8 nnt.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Oxidizing  (fuel lean).
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:   Dissolve  1.303  g  thallium nitrate, T1N03
     (analytical  reagent  grade),  in   Type   II  water,  acidify  with  10 mL
     concentrated HN03, and dilute to  1  liter  with Type II water.  Alterna-
     tively, procure  a  certified  standard   from  a  supplier  and verify by
     comparison with a second standard.
                                  7840 - 1
                                                         Revision
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution   to   be   used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  acid and   at   the  same
     concentration  as  will   result  1n  the  sample  to  be   analyzed after
     processing (0.5% v/v
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The  performance  characteristics  for  an  aqueous  sample  free of
interferences are:

     Optimum concentration range:  1-20 mg/L with a wavelength of 276.8 nm.
     Sensitivity:  0.5 mg/L.
     Detection limit:  0.1 mg/L.

     9.2  In a   single   laboratory,  analysis  of  a mixed  Industrial -domestic
waste effluent,  digested with Method 3010, at concentrations of 0.6, 3, and 15
mg/L gave  standard  deviations  of  +0.018,  +0.05,  and   +0.2, respectively.
Recoveries at these  levels were  100%,  98%, and~98%, respectively.

     9.3  For concentrations of  thallium below 0.2 mg/L, the furnace technique
 (Method 7841) is recommended.


 10.0 REFERENCES

 1.   Methods for  Chemical   Analysis  of   Water   and Wastes, EPA-600/4-82-055,
 December  1982, Method  279.1.
                                   7840 - 2
                                                          Revision
                                                          Date  September 1986

-------
                 METHOD 7840

THALLIUM  (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
                5.0
                   Prepare
                  standards
7. 1
prepar
cl-
sec
For
sample
•atlorv see
\apter 3.
:tion 3.2

7.2


Analyze using
Method 70OO.
Section 7.2
             f     Stop      J
            7840 -  3
                                     Revision      0
                                     Date  September 1986

-------
                                 METHOD 7841

               THALLIUM (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction 1s required.

     3.3  Hydrochloric add or excessive chloride will cause volatilization of
thallium at low temperatures.  Verification  that losses are not occurring, by
spiked samples or standard additions, must be made for each sample matrix.

     3.4  Palladium is a suitable matrix modifier for thallium analysis.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Drying time and temp:  30 sec at 125*C.
          4.2.2  Ashing time and temp:  30 sec at 400*C.
          4.2.3  Atomizing time and temp:   10 sec at 2400*C.
          4.2.4  Purge gas:  Argon or nitrogen.
          4.2.5  Wavelength:  276.8 nm.
          4.2.6  Background correction:   Required.
          4.2.7  Other operating parameters should be  set as specified by the
                 particular instrument manufacturer.
     NOTE:  The above concentration values and instrument conditions are for a
            Perkin-Elmer HGA-2100, based  on  the  use  of  a 20-uL injection,
            continuous-flow purge  gas,  and  nonpyrolytic  graphite.  Smaller
            sizes of  furnace  devices  or  those  employing  faster  rates of
            atomization can be  operated  using lower atomization temperatures
            for shorter time periods than the above-recommended settings.
                                  7841 - 1
                                                         Revision
                                                         Date  September 1986

-------
5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:   Dissolve  1.303  g  thallium nitrate, T1N03
     (analytical  reagent  grade),  1n  Type  II  water,  acidify  with  10 ml
     concentrated HNOs, and dilute to  1  liter  with Type II water.  Alterna-
     tively, procure  a  certified  standard  from  a  supplier  and verify by
     comparison with a second standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  acid  and  at  the same
     concentrations as 1n the sample after processing (0.5% v/v
     5.3  Palladium chloride:  Weigh 0.25 g  of PdCl2 to the nearest 0.0001 g.
Dissolve 1n  10 ml of  1:1 HNOs and  dilute  to 1 liter with Type II water.  Use
equal volumes of sample and palladium solution.
6.0  SAMPLE  COLLECTION,  PRESERVATION, AND HANDLING

     6.1   See  Chapter  Three,  Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1   Sample preparation;   The  procedures  for  preparation of the sample
are  given  1n Chapter Three,  Section 3.2.

     7.2   See  Method 7000,  Paragraph 7.3,   Furnace Procedure.  The calculation
1s given  1n  Method 7000,  Paragraph 7.4.


8.0  QUALITY CONTROL

     8.1   See  Section  8.0 of Method 7000.


9.0  METHOD  PERFORMANCE

     9.1   Precision and accuracy data are not available  at this time.

     9.2   The   performance  characteristics for  an   aqueous  sample  free of
Interferences  are:

     Optimum concentration range: 5-100 ug/L.
     Detection limit:   1  ug/L.
                                   7841 - 2
                                                          Revision
                                                          Date  September 1986

-------
10.0 REFERENCES

1.   Application  of  Matrix-Modification  1n  Determination  of  Thallium  in
Wastewater by Graphite-Furnace  Atomic-Absorption Spectrometry,  Talanta, 31(2)
(1984), pp. 150-152.
                                   7841 - 3
                                                          Revision
                                                          Date   September  1986

-------
                METHOD 7841

THALLIUM  (ATOMIC ABSORPTION.  FURNACE TECHNIQUE)
5.0

Prepare
standards

7. 1
prepar
ct-
sec

7.2


For
sample
atlon see
apter 3.
tlon 3.2



Analyze using
Method 700O.
Section 7.2:
Calculation 7.4
             f    Stop      J
            7841 - 4
                                    Revision      0
                                    Date  September  1986

-------
                                 METHOD 7870

                 TIN (ATOMIC ABSORPTION, DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Tin hollow cathode lamp.
          4.2.2  Wavelength:  286.3 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxidant:  Nitrous oxide.
          4.2.5  Type of flame:  Fuel rich.
          4.2.6  Background correction:  Not required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5«2  Preparation of standards;

          5.2.1  Stock solution:  Dissolve  1.000  g  of tin metal  (analytical
     reagent grade) in 100 mL of  concentrated  HC1 and dilute to 1 liter with
     Type II  water.    Alternatively,  procure  a  certified  standard from a
     supplier and verify by comparison  with a second standard.

          5.2.2  Prepare  dilutions  of the  stock  solution  to  be  used as
     calibration standards at the time  of analysis.  The calibration standards
     should  be  prepared  using  the   same  type  of  acid  and  at  the same
     concentration  as  will  result  in  the  sample  to  be  analyzed  after
     processing.
                                  7870 - 1
                                                         Revision
                                                         Date  September 1986

-------
6.0  SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

     6.1  See Chapter Three, Section 3.1.3,  Sample Handling and Preservation.

7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  The  performance  characteristics  for  an  aqueous  sample  free of
interferences are:

     Optimum concentration  range:   10-300 mg/L with a wavelength of 286.3 nm.
     Sensitivity:  4 mg/L.
     Detection  limit:  0.8  mg/L.

     9.2   In a   single  laboratory,   analysis  of  a mixed  industrial-domestic
waste  effluent,  digested with  Method  3010,  at concentrations  of 4, 20, and 60
mg/L   gave   standard   deviations   of   +0.25,  +0.5,  and  +0.5,  respectively.
Recoveries  at these  levels  were 96%,  101%,  and 101%, respectively.


10.0 REFERENCES

1.   Methods for  Chemical   Analysis   of Water   and Wastes, EPA-600/4-82-055,
December  1982,  Method  282.1.
                                   7870 - 2
                                                          Revision
                                                          Date  September 1986

-------
               METHOD 7B7O

TIN (ATOMIC  ABSORPTION.  DIRECT ASPIRATION)
s.
O

Prepare
standards
               7.1
                   1 For
                   sample
              preparation see
                 chapter 3.
                 section 3.2
               7.2
               Analyze using
               Method 7000.
               Section 7.2
            f     Stop      J
           7870 -  3
                                     Revision       0
                                     Date   September 1986

-------
                                 METHOD 7910

               VANADIUM (ATOMIC ABSORPTION,  DIRECT ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction may be required.

     3.3  High concentrations of aluminum or  titanium, or the presence of Bi,
Cr, Co, Fe, acetic acid,  phosphoric  acid, surfactants, detergents, or alkali
metals,  may  interfere.    The  interference  can  be  controlled  by  adding
1,000 mg/L aluminum to samples and standards.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Vanadium hollow cathode lamp.
          4.2.2  Wavelength:   318.4 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:   Nitrous oxide.
          4.2.5  Type of  flame:  Fuel  rich.
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of  standards;

          5.2.1  Stock solution:   Dissolve   1.7854  g  of vanadium pentoxide,
     ¥205  (analytical reagent  grade),  in 10 mL of concentrated nitric acid and
     dilute to 1 liter with Type II water.  Alternatively, procure a certified
     standard from a supplier  and verify by comparison with a second standard.
                                   7910 -  1
                                                         Revision
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution   to   be   used  as
     calibration standards at the time of analysis.  The  calibration standards
     should  be  prepared  using  the  same  type   of  add  and   at   the  same
     concentration  as  will   result  1n  the  sample  to be   analyzed after
     processing.  In addition, 2 ml of the aluminum nitrate  solution described
     In Paragraph 5.2.3  should  be  added  to  each  100 ml  of standards and
     samples.

          5.2.3  Aluminum nitrate solution:   Dissolve  139  g  aluminum nitrate
     (A1[N03]3«9H20) In 150 ml  Type  II  water; heat to  complete  dissolution.
     Allow to cool and dilute to 200  ml  with Type II water.   All  samples and
     standards should contain 2 ml of this solution per 100  ml.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three,  Section 3.1.3, Sample Handling  and Preservation.

7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD  PERFORMANCE

     9.1  The   performance  characteristics  for   an  aqueous  sample  free of
Interferences are:

     Optimum concentration range:   2-100 mg/L with  a wavelength of 318.4 nm.
     Sensitivity:   0.8 mg/L.
     Detection  limit:  0.2 mg/L.

     9.2   In a  single  laboratory),   analysis   of   a mixed  Industrial-domestic
waste  effluent,  digested with  Method  3010,   at  concentrations  of 2, 10, and 50
mg/L   gave   standard   deviations   of    +0.1,  +0.1,  and  +0.2,  respectively.
Recoveries at  these levels were 100%,  95%,  and  97%,  respectively.

     9.3   For  concentrations  of vanadium below  0.5 mg/L, the  furnace  technique
 (Method 7911)  1s recommended.
                                   7910 -  2
                                                         Revision
                                                         Date  September 1986

-------
10.0 REFERENCES

1.   Methods for  Chemical  Analysis  of  Water  and Wastes, EPA-600/4-82-055,
December 1982, Method 286.1.
                                  7910 - 3
                                                         Revision
                                                         Date  September 1986

-------
                 METHOD 791O

VANADIUM (ATOMIC ABSORPTION.  DIRECT ASPIRATION)
                s.o
                   Prepare
                   standards
7.1
prepar
ch
sec
For
sample
atlon see
apter 3.
tlon 3. a
               7.2
               Analyze using
                Method 7000.
                Section 7.Z
             (     Stop      J
               7910
                                         Revision       0
                                         Date   September 1986

-------
                                 METHOD 7911

               VANADIUM (ATOMIC ABSORPTION.  FURNACE  TECHNIQUE)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  Background correction is required.

     3.3  Vanadium is refractory and  prone  to  form carbides.  Consequently,
memory effects are common,  and  care  should  be  taken  to clean the furnace
before and after analysis.

     3.4  Nitrogen should not be used as a purge gas.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters (general):

          4.2.1  Drying time and temp:  30 sec at 125*C.
          4.2.2  Ashing time and temp:  30 sec at 1400*C.
          4.2.3  Atomizing time and temp:  15 sec at 2800*C.
          4.2.4  Purge gas:  Argon (nitrogen should not be used).
          4.2.5  Wavelength:  318.4 nm.
          4.2.6  Background correction:   Required.
          4.2.7  Other operating parameters should be  set as specified by the
                 particular instrument manufacturer.
     NOTE:  The above concentration values and instrument conditions are for a
            Perkin-Elmer HGA-2100, based  on  the  use  of  a 20-uL injection,
            continuous-flow purge  gas,  and  nonpyrolytic  graphite.  Smaller
            sizes of  furnace  devices  or  those  employing  faster  rates of
            atomization can be  operated  using lower atomization temperatures
            for shorter time periods than the above-recommended settings.
                                  7911 - 1
                                                         Revision
                                                         Date  September 1986

-------
5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:   Dissolve  1.7854  g  of vanadium pentoxlde,
     V2®5 (analytical reagent grade), 1n 10 mL of concentrated nitric add and
     dilute to 1 liter with Type II water.  Alternatively, procure a certified
     standard from a supplier and verify by comparison with a second standard.

          5.2.2  Prepare  dilutions  of  the  stock  solution  to  be  used as
     calibration standards at the time of analysis.  The calibration standards
     should  be  prepared  using  the  same  type  of  acid  and  at  the same
     concentrations as in the sample after processing (0.5% v/v
6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample  preparation;   The  procedures  for  preparation of the sample
are given 1n Chapter Three, Section 3.2.

     7.2  See Method 7000,  Paragraph 7.3,   Furnace Procedure.  The calculation
is given in Method  7000,  Paragraph 7.4.


8.0  QUALITY CONTROL

     8.1  See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

     9.1  Precision and accuracy data are available  1n Method 286.2 of Methods
for  Chemical Analysis of Water and Wastes.

     9.2  The   performance characteristics  for  an aqueous  sample  free  of
interferences are;

     Optimum concentration range:  10-200 ug/L.
     Detection  limit:   4 ug/L.


 10.0 REFERENCES

 1.    Methods  for  Chemical  Analysis   of Water  and Wastes,  EPA-600/4-82-055,
 December 1982,  Method 286.2.


                                   7911 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
                 METHOD 7911

VANADIUM  (ATOMIC ABSORPTION.  FURNACE  TECHNIQUE)
                5.O
                   Prepare
                   standards
7. 1
f
• I
preparat
chat
sect!

•or
imple
.Ion see
>ter 3.
ion .3.2

7.2 |
Analyze using
Method 70OO.
Section 7.2:
calculation 7.4
              f     Stop      J
            7911 - 3
                                     Revision       0
                                     Date  September  1986

-------
                                 METHOD 7950

                 ZINC (ATOMIC ABSORPTION,  DIRECT  ASPIRATION)


1.0  SCOPE AND APPLICATION

     1.1  See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

     2.1  See Section 2.0 of Method 7000.


3.0  INTERFERENCES

     3.1  See Section 3.0 of Method 7000 if interferences are suspected.

     3.2  High levels of silicon, copper,   or  phosphate may interfere.  Addi-
tion of strontium (1,500 mg/L) removes the copper and phosphate interference.

     3.3  Zinc is a universal contaminant,  and  great care should be taken to
avoid contamination.


4.0  APPARATUS AND MATERIALS

     4.1  For basic apparatus, see Section 4.0 of Method 7000.

     4.2  Instrument parameters  (general):

          4.2.1  Zinc hollow cathode lamp.
          4.2.2  Wavelength:  213.9 nm.
          4.2.3  Fuel:  Acetylene.
          4.2.4  Oxldant:  Air.
          4.2.5  Type of flame:  Oxidizing (fuel  lean).
          4.2.6  Background correction:  Required.


5.0  REAGENTS

     5.1  See Section 5.0 of Method 7000.

     5.2  Preparation of standards;

          5.2.1  Stock solution:    Dissolve  1.000  g  zinc metal (analytical
     reagent grade) in 10 ml of concentrated nitric acid and dilute to 1 liter
     with Type II water.   Alternatively,  procure a certified standard from a
     supplier and verify by comparison with a second standard.
                                  7950 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
          5.2.2  Prepare  dilutions  of  the  stock  solution   to   be   used  as
     calibration standards at the time of analysis.  The calibration standards
     should be prepared using the same type of acid and at  the same concentra-
     tion as will  result in the sample to be analyzed after processing.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See Chapter Three, Section 3.1.3, Sample Handling and Preservation.


7.0  PROCEDURE

     7.1  Sample preparation;  The  procedures  for  preparation of the sample
are given in Chapter Three, Section 3.2.

     7.2  See Method 7000, Paragraph 7.2, Direct Aspiration.


8.0  QUALITY CONTROL

     8.1  See Section  8.0 of Method  7000.


9.0  METHOD  PERFORMANCE

     9.1  The   performance   characteristics   for   an   aqueous  sample  free of
interferences are:

     Optimum concentration  range:   0.05-1  mg/L with  a  wavelength of 213.9 nm.
     Sensitivity:   0.02 mg/L.
     Detection  limit:   0.005 mg/L.

     9.2  For concentrations of zinc  below  0.01  mg/L, the  furnace technique
 (Method 7951)  is  recommended.

     9.3  Precision and accuracy data are  available  in Method 289.1 of Methods
for  Chemical  Analysis  of Water and Wastes.


 10.0 REFERENCES

 1.   Methods for   Chemical   Analysis  of  Water   and Wastes,  EPA-600/4-82-055,
December 1982,  Method  289.1.
                                   7950 - 2
                                                          Revision
                                                          Date  September 1986

-------
               METHOD 7950

ZINC (ATOMIC  ABSORPTION.  OIRCCT ASPIRATION)
              5.0
                  Prepare
                 standards
              7.1
                    For
                   sample
             preparation see
                 chapter 3.
                section 3.2
              7 .Z
              Analyze using
               Method 7OOO.
               Section 7.2
            f     Stop      J
             7950 - 3
                                       Revision       0
                                       Date  September 1986

-------
                                  METHOD  7951

                  ZINC  (ATOMIC  ABSORPTION,  FURNACE TECHNIQUE)


1.0  SCOPE AND APPLICATION

      1.1   See Section 1.0 of Method 7000.


2.0  SUMMARY OF METHOD

      2.1   See Section 2.0 of Method 7000.


3.0  INTERFERENCES

      3.1   See Section 3.0 of Method 7000.

      3.2   Background correction should be used.

      3.3   Zinc  is  a  universal  contaminant.   Because  of this and  the high
sensitivity of this method, great care should be taken to avoid contamination.


4.0  APPARATUS AND MATERIALS

      4.1   For basic apparatus, see Section 4.0 of Method 7000.

      4.2   Instrument parameters (general):

            4.2.1  Drying time and temp:  30 sec at 125°C.

            4.2.2  Ashing time and temp:  30 sec at 400°C.

            4.2.3  Atomizing time and temp:  10 sec at 2500°C.

            4.2.4  Purge gas: Argon or nitrogen.

            4.2.5  Wavelength:  213.9 nm.

            4.2.6  Background correction:  Required.

            4.2.7  Other operating parameters should be  set as specified by the
      particular instrument manufacturer.

            NOTE: The above concentration values and instrument conditions are
                  for  a  Perkin-Elmer HGA-2100,  based  on  the  use of  a 20-uL
                  injection,   continuous-flow   purge  gas,  and  nonpyrolytic
                  graphite.   Smaller  size furnace  devices or  those employing
                  faster  rates of  atomization  can be  operated using  lower


                                   7951 - 1                       Revision 0
                                                                  July 1992

-------
                  atomization temperatures  for  shorter time periods  than the
                  above-recommended settings.


5.0  REAGENTS

      5.1   See Section 5.0 of Method 7000.

      5.2   Preparation of standards

            5.2.1  Stock  solution  - Dissolve 1.000  g zinc  metal  (analytical
      reagent grade) in 10 mL of concentrated nitric acid and dilute to 1 liter
      with water.  Alternatively, procure a certified standard from a supplier
      and verify by comparison with a second standard.

            5.2.2  Prepare  dilutions of  the stock   solution  to  be used  as
      calibration standards at the  time  of  analysis.  The calibration standards
      should  be  prepared  using  the  same  type  of  acid   and  at  the  same
      concentrations as in the sample after processing (0.5% v/v HN03).


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See Chapter Three, Step 3.1.3,  Sample Handling and Preservation.


7.0  PROCEDURE

      7.1   Sample Preparation -  The procedures for preparation of the sample are
given in Chapter Three, Step 3.2.

      7.2   See Method 7000, Step 7.3, Furnace Technique.


8.0  QUALITY CONTROL

      8.1   See Section 8.0 of Method 7000.


9.0  METHOD PERFORMANCE

      9.1   Precision and  accuracy data are  not available at this time.

      9.2   The  performance  characteristics for  an   aqueous  sample  free  of
interferences are:

      Optimum concentration range:  0.2-4  ug/L.
      Detection limit:  0.05 ug/L.
                                   7951 - 2                       Revision 0
                                                                  July  1992

-------
10.0  REFERENCES

1.    Methods for  Chemical  Analysis of  Water  and Wastes;  U.S.  Environmental
Protection Agency.  Office of Research and Development. Environmental Monitoring
and Support  Laboratory.  ORD Publication Offices  of Center  for  Environmental
Research Information:  Cincinnati, OH,  1983; EPA-600/4-79-020.
                                   7951 - 3                       Revision 0
                                                                  July 1992

-------
                   METHOD  7951
ZINC  (ATOMIC ABSORPTION,  FURNACE TECHNIQUE)
                                                                        i
                    Start
                 5.0 Pr«par«
                  •tandard*
                7.1 For «ampl«
                preparation •••
              Chapter 3, Section
                     3.2
               7.2 Analyz* using
                M.thod 7000
                 Section 7.3
                    Stop
                 7951  -  4
Revision  0
July 1992

-------
                                  APPENDIX

                             COMPANY REFERENCES


The  following  listing  of frequently-used  addresses  Is  provided  for  the
convenience of users of this manual.   No endorsement  1s Intended or Implied.


Ace Glass Company
1342 N.W. Boulevard
P.O. Box 688
Vlneland, NJ  08360
(609) 692-3333

Aldrlch Chemical Company
Department T
P.O. Box 355
Milwaukee, WI  53201

Alpha Products
5570 - T W. 70th Place
Chicago, IL  60638
(312) 586-9810

Barneby and Cheney Company
E. 8th Avenue and N. Cassldy Street
P.O. Box 2526
Columbus, OH  43219
(614) 258-9501

Bio - Rad Laboratories
2200 Wright Avenue
Richmond, CA  94804
(415) 234-4130

Burdick & Jackson Lab Inc.
1953 S. Harvey Street
Muskegon, MO  49442

Calgon Corporation
P.O. Box 717
Pittsburgh, PA  15230
(412) 777-8000

Conostan Division
Conoco Speciality Products, Inc.
P.O. Box 1267
Ponca City, OK  74601
(405) 767-3456
                                COMPANIES -  1
                                                         Revision
                                                         Date  September  1986

-------
Corning Glass Works
Houghton Park
Corning, NY  14830
(315) 974-9000

Dohrmann, Division of Xertex Corporation
3240 - T Scott Boulevard
Santa Clara, CA  95050
(408) 727-6000
(800) 538-7708

E. M. Laboratories, Inc.
500 Executive Boulevard
Elmsford, NY  10523

Fisher Scientific Co.
203 Fisher Building
Pittsburgh, PA  15219
(412) 562-8300

General Electric Corporation
3135 Easton Turnpike
Fairfield, CT  06431
(203) 373-2211

Graham Manufactory Co., Inc.
20 Florence Avenue
Batavia, NY  14020
(716) 343-2216

Hamilton Industries
1316 18th Street
Two  Rivers, WI  54241
(414) 793-1121

ICN  Life Sciences Group
3300 Hyland Avenue
Costa Mesa, CA  92626

Johns - Manvllle Corporation
P.O. Box 5108
Denver,  CO  80217

Kontes  Glass Company
8000 Spruce Street
Vineland,  NJ  08360

Millipore  Corporation
80 Ashby Road
Bedford, MA 01730
 (617)  275-9200
 (800)  225-1380
                                 COMPANIES - 2
                                                          Revision       0
                                                          Date   September  1986

-------