SW846PU32
                     DETERMINATION OF CARBONYL COMPOUNDS
               BY HIGH  PERFORMANCE  LIQUID CHROMATOGRAPHY (HPLC)
1.0  SCOPE AND APPLICATION

     1.1  This method provides procedures for the determination of free carbonyl
compounds in  various matrices by derivatization with 2,4-dinitrophenylhydrazine
(DNPH).   The  method utilizes high performance liquid chromatography (HPLC) with
ultraviolet/visible  (UV/vis)  detection to  identify and  quantitate  the target
analytes.  This method includes  two  procedures encompassing all aspects of the
analysis  (extraction  to  determination  of  concentration).    Procedure  1  is
appropriate for the analysis of aqueous, soil and waste samples and stack samples
collected by  Method 0011.  Procedure  2  is appropriate for the  analysis of indoor
air samples collected by Method 0100.   The list of target analytes differs by
procedure. The appropriate procedure  for  each target analyte is listed in the
table below.
      Compound                              CAS No."       Proc. lb   Proc.  2
                                                                            b
Acetaldehyde
Acetone
Acrolein
Benzaldehyde
Butanal (butyraldehyde)
Crotonaldehyde
Cyclohexanone
Decanal
2,5-Dimethylbenzaldehyde
Formaldehyde
Heptanal
Hexanal (hexaldehyde)
Isovaleraldehyde
Nonanal
Octanal
Pentanal (valeraldehyde)
Propanal (propionaldehyde)
m-Tolualdehyde
o-Tolualdehyde
p-Tolualdehyde
75-07-0
67-64-1
107-02-8
100-52-7
123-72-8
123-73-9
108-94-1
112-31-2
5779-94-2
50-00-0
111-71-7
66-25-1
590-86-3
124-19-6
124-13-0
110-62-3
123-38-6
620-23-5
529-20-4
104-87-0
X



X
X
X
X

X
X
X

X
X
X
X
X


X
X
X
X
X
X


X
X

X
X


X
X
X
X
X
         Chemical Abstract Services Registry Number.

         The two procedures have overlapping lists of target compounds that have
         been evaluated  using  modifications of the  analysis.   Refer  to  the
         respective procedure  number  when choosing  the  appropriate  analysis
         technique for a particular compound.
                                  8315A -  1                         Revision 1
                                                                 January 1995

-------
      1.2  Method detection limits (MDL)  using procedure 1 are listed in Tables 1
and  2.   Sensitivity  data for  sampling and  analysis  by  use  of Method  0100
(procedure 2) are given  in Table 3.   The MDL  for  a  specific  sample  may differ
from the listed  value, depending upon the interferences  in the sample matrix and
the volume of sample used in the procedure.

      1.3  The extraction procedure for solid samples is outlined  in Sec. 7.1 of
this method.

      1.4  When  this  method  is  used  to analyze  unfamiliar  sample  matrices,
compound  identification  should  be   supported  by  at  least  one  additional
qualitative technique. A gas  chromatograph/mass spectrometer (GC/MS) may be used
for the qualitative confirmation of results for the  target analytes,  using the
extract produced by this method.

      1.5  This  method is restricted  to  use by,  or under  the  supervision of,
analysts experienced in the use of chromatography and in the  interpretation of
chromatograms.   Each analyst must demonstrate the ability to generate acceptable
results with this method.
2.0  SUMMARY OF  METHOD

     This method contains two procedures dealing with different sample types.

     2.1  Liquid and  Solid  Samples  (Procedure  1)

          2.1.1    For wastes comprised of solids,  and aqueous wastes containing
     greater  than   one  percent solid  material,  the aqueous  phase  should  be
     separated  from the solid phase  and  stored,  according to Sec.  6.2,   for
     possible later analysis.  If necessary,  the particle size of the solids in
     the waste is reduced.  The solid phase is  extracted with a volume of fluid
     equal  to 20 times  the  sample's weight.   The  extraction fluid employed is
     a  function  of  the  alkalinity of the  solid phase of the waste.  A special
     extractor   is  used  when  volatiles  are  being  extracted.    Following
     extraction, the  extract is  filtered  through  a  0.6 -  0.8 /zm glass fiber
     filter.

          2.1.2    If  compatible   (i.e.,  multiple  phases   will   not  form  on
     combination),  the initial  aqueous phase of the  waste  is  added  to  the
     aqueous extract, and these liquids  are analyzed together.  If incompatible,
     the  liquids are analyzed  separately  and the results  are  mathematically
     combined to yield  a  volume-weighted average  concentration.

          2.1.3   A measured volume of aqueous sample  (approx.  100 mL) or an
     appropriate amount of  solids extract  (approx.  25  g),  is buffered to pH 3
     and  derivatized  with 2,4-dinitrophenylhydrazine  (DNPH),  using either the
     appropriate solid-phase or  a liquid-liquid  extraction technique.  If the
     solid-phase extraction (SPE) option  is  used, the  derivatized compound is
     extracted using solid sorbent cartridges,  then eluted with ethanol.  If the
     liquid-liquid  option   is  used,  the  derivatized  compound   is  serially
     extracted three  (3) times with methylene chloride.  The methylene chloride
     extracts  are  concentrated  using  the  appropriate  procedure  3500  series
     method and exchanged  with  acetonitrile  prior to  HPLC  analysis.   HPLC

                                   8315A -  2                         Revision 1
                                                                  January 1995

-------
     conditions  are described which permit the separation  and  measurement of
     various  carbonyl  compounds  in  the  extract  by absorbance detection at 360
     nm.

           2.1.4    If formaldehyde is the only analyte of interest, the aqueous
     sample and/or solid sample extract  should be buffered to pH  5.0 to minimize
     the  formation  of artifact formaldehyde.

     2.2   Stack  Gas  Samples Collected by Method  0011  (Procedure  1):  The entire
sample returned to the laboratory is extracted with methylene chloride and the
extract  is diluted  or concentrated to  a  known  volume.    An  aliquot  of the
methylene chloride extract is solvent exchanged and concentrated or diluted as
necessary.    HPLC  conditions  are  described  that permit  the   separation  and
measurement of various carbonyl compounds in the  extract by absorbance detection
at 360 nm.

     2.3   Indoor Air  Samples by   Method  0100  (Procedure  2):    The  sample
cartridges are returned to the laboratory and backflushed with acetonitrile into
a 5-mL  volumetric flask.  The eluate  is diluted to volume with acetonitrile.  Two
aliquots of the acetonitrile extract are pipetted into two sample vials having
Teflon®-!ined  septa.   HPLC  conditions are  described  which   allow   for  the
separation and measurement of the various carbonyl compounds in the extract by
absorbance detection at 360 nm.
3.0   INTERFERENCES

      3.1   Method interferences may be caused by contaminated solvents, reagents,
glassware,  and  other sample  processing  hardware which  can  lead  to  discrete
artifacts and/or elevated chromatogram baselines.  All  materials should routinely
demonstrate freedom  from  interferences under  analysis  conditions by analyzing
laboratory reagent blanks as described in Sec. 8.5.

           3.1.1    Glassware must  be scrupulously cleaned.  Glassware should be
      rinsed  as  soon as possible  after use  with  the last solvent  used.   This
      should be followed by detergent washing with hot water, and rinses with tap
      water and organic-free reagent water.  After washing the glassware should
      then  be  drained,  dried,  and  heated  in  a  laboratory oven at  130"C for two
      to  three hours before reuse.  Solvent  rinses  with acetonitrile  may be
      substituted  for the oven  heating.   After drying  and cooling,  glassware
      should be stored in a clean environment to prevent  accumulation of dust or
      other contaminants.

NOTE: Do not  rinse  glassware  with acetone or  methanol.   These  solvents react
      with DNPH to form interferences.

           3.1.2   The  use of  high purity reagents and  solvents  helps minimize
      interference.   Purification of solvents by  distillation  in all  glass
      systems  may be  required.

           3.1.3    Polyethylene  gloves must be worn  when handling  silica gel
      cartridges to reduce the possibility of  contamination.
                                   8315A  - 3                         Revision 1
                                                                  January 1995

-------
      3.2   Formaldehyde   contamination   of  the  DNPH  reagent  is  frequently
encountered due  to its  widespread  occurrence in  the  environment.    The  DNPH
reagent in Procedure 2 must be  purified  by multiple recrystallizations in HPLC-
grade acetonitrile.   Recrystallization  is accomplished,  at  40 -  60°C, by slow
evaporation of the solvent  to maximize crystal size.  The  purified DNPH crystals
are stored under  HPLC-grade acetonitrile.  Impurity levels of carbonyl compounds
in the DNPH are determined prior to sample analysis and should be less than 25
mg/L.  Refer to Appendix A for the recrystallization procedure.

      3.3   Matrix interferences may be caused by contaminants co-extracted from
the  sample.   The extent of matrix  interferences will  be source-  and matrix-
specific.   If  interferences occur in subsequent  samples, modification  of the
mobile phase or some additional cleanup may be necessary.

      3.4   In  Procedure  1,  acetaldehyde  is generated during  the derivatization
step  if  ethanol  is present  in the sample.   This background  will  impair the
measurement of acetaldehyde levels below 0.5 ppm (500 ppb).

      3.5   For  Procedure 2, at the stated two  column analysis conditions, the
identification  and quantitation  of  butyraldehyde  may  be  difficult due  to
coelution with isobutyraldehyde and methyl ethyl  ketone.  Precautions  should be
taken and adjustment of the analysis conditions  should be made to avoid potential
problems.


4.0   APPARATUS AND MATERIALS

      4.1   High performance liquid  chromatograph  (modular).

           4.1.1    Pumping  system - Gradient, with constant flow control capable
      of  1.50 mL/min.

           4.1.2    High  pressure  injection valve  with 20-juL  loop.

           4.1.3    Column - 250 mm x 4.6 mm ID, 5-jiim  particle size, CIS (Zorbax
      or  equivalent).

           4.1.4    Absorbance  detector  - 360  nm.

           4.1.5    Strip-chart  recorder  compatible with detector -  Use  of a data
      system for  measuring  peak areas and  retention times is  recommended.

           4.1.6    Helium  - for  degassing mobile  phase  solvents.  (Procedures
      1 and 2)

           4.1.7    Mobile phase reservoirs and suction filtration apparatus  - For
      holding  and filtering HPLC  mobile  phase.   Filtering system should be all
      glass and Teflon®  and use a  0.22 ;um polyester membrane filter.

           4.1.8    Syringes -  for  HPLC injection  loop loading, with capacity at
      least four  times the  loop volume.
                                   8315A -  4                        Revision 1
                                                                  January 1995

-------
4.2  Apparatus and materials for Procedure 1
     4.2.1   Reaction vessel - 250-mL Florence flask.
     4.2.2   Separatory funnel - 250-mL, with Teflon® stopcock.
     4.2.3   Kuderna-Danish  (K-D)  apparatus - See  Method 3510 or  other
appropriate 3500  series  method.    (Other  concentration  apparatus may  be
employed if the laboratory can demonstrate equivalent performance).
     4.2.4   Boiling chips  - Solvent-extracted with methylene  chloride,
approximately 10/40 mesh (silicon carbide or equivalent).
     4.2.5   pH meter - Capable of measuring to 0.01 pH units.
     4.2.6   Glass fiber filter paper -  1.2 /zm pore size (Fisher Grade  G4
or equivalent).
     4.2.7   Solid  sorbent  cartridges -  Packed  with 2  g CIS  (Baker  or
equivalent).
     4.2.8   Vacuum manifold -  Capable of  simultaneous extraction of up  to
12 samples (Supelco or equivalent).
     4.2.9   Sample reservoirs - 60-mL capacity (Supelco  or equivalent).
     4.2.10  Pipet - Capable of accurately delivering 0.10 ml solution.
     4.2.11  Water bath - Heated, with concentric  ring  cover,  capable  of
temperature control (± 2°C).  The bath should be used in a hood.
     4.2.12  Sample shaker - Controlled temperature incubator (± 2°C) with
orbital shaking (Lab-Line Orbit Environ-Shaker Model  3527 or equivalent).
     4.2.13  Syringes - 5-mL, 500-/iL, 100-^L,  (Luer-Lok or equivalent).
     4.2.14  Syringe filters - 0.45 p.m filtration  disks (Gelman Acrodisc
4438 or equivalent).
4.3  Apparatus and materials for Procedure 2
     4.3.1   Syringes  -  10-mL,  with Luer-Lok  type  adapter,  used  to
backflush the sample cartridges by gravity feed.
     4.3.2   Syringe rack -  made of an aluminum plate with adjustable legs
on all  four corners.  Circular holes of a diameter slightly larger than the
diameter of the 10-mL syringes are drilled  through the plate to allow batch
processing of  cartridges  for cleaning,  coating, and  sample  elution.   A
plate (0.16 x 36  x 53 cm)  with  45  holes  in a 5 x  9  matrix is recommended.
See Figure 2 in Method 0100.
4.4  Volumetric flasks -  5-mL, 10-mL, and 250- or 500-mL.
                             8315A - 5                         Revision  1
                                                             January  1995

-------
     4.5   Vials  -  10- or 25-mL, glass with Teflon®-lined  screw caps or crimp
tops.

     4.6   Balance  - Analytical,  capable  of accurately weighing  to  0.0001  g.

     4.7   Glass  funnels

     4.8   Polyethylene gloves  -  used  to  handle  silica gel  cartridges.


5.0  REAGENTS

     5.1   Reagent  grade inorganic chemicals shall be used in  all tests.  Unless
otherwise  indicated,  it  is  intended  that all  reagents  shall  conform  to the
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where such  specifications  are available.  Other  grades  may be used,
provided it is  first ascertained  that the reagent is  of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

     5.2   Organic-free reagent water -  Water  in which  an interferant is not
observed at the method detection limit for the compounds of interest.

     5.3   Formalin -  Solution of formaldehyde  (CH20)  in organic-free reagent
water,  nominally 37.6  percent (w/w).  Exact concentration  will  be determined for
the stock solution in Sec. 5.7.1.1.

     5.4   Aldehydes and ketones - analytical grade, used for preparation of DNPH
derivative standards of target analytes other than formaldehyde.   Refer to the
target analyte list.

     5.5   Procedure  1 reagents

           5.5.1    Methylene  chloride,  CH2C12  - HPLC grade  or equivalent.

           5.5.2    Acetonitrile,  CH3CN  -  HPLC grade or equivalent.

           5.5.3    Sodium  hydroxide  solutions, NaOH, 1.0  N and 5 N.

           5.5.4    Sodium   chloride,  NaCl,  saturated   solution  -  Prepare  by
     dissolving  an excess of reagent grade solid in organic-free reagent water.

           5.5.5    Sodium  sulfite solution, Na2S03, 0.1 M.

           5.5.6    Sodium  sulfate, Na2S04  - granular, anhydrous.

           5.5.7    Citric  acid, C8H807, 1.0 M solution.

           5.5.8    Sodium  citrate, C6H5Na307«2H20, 1.0 M trisodium salt dihydrate
     solution.

           5.5.9    Acetic  acid (glacial), CH3C02H.

           5.5.10  Sodium  acetate, CH3C02Na.


                                  8315A  - 6                         Revision  1
                                                                   January 1995

-------
     5.5.11  Hydrochloric acid, HC1, 0.1 N.

     5.5.12  Citrate buffer,  1 M, pH 3   -  Prepare  by adding 80 ml of 1 M
citric  acid  solution  to 20  mL  of  1  M  sodium citrate  solution.   Mix
thoroughly.  Adjust pH with NaOH or HC1 as needed.

     5.5.13  pH  5.0 Acetate  buffer (5M)  -  Formaldehyde  analysis only.
Prepared by adding 40 ml  5M acetic acid solution to  60 ml 5M sodium acetate
solution.  Mix thoroughly.  Adjust pH with NaOH or HC1 as needed.

     5.5.14  2,4-Dinitrophenylhydrazine, 2,4-(02N)2C6H3]NHNH2, (DNPH), 70%
in organic-free  reagent  water (w/w).   Prepare a 3.00 mg/mL  solution by
dissolving 428.7 mg of 70% (w/w)  DNPH solution in 100 ml of acetonitrile.

     5.5.15  Extraction  fluid for Procedure  1  -  Dilute  64.3  mL of 1.0 N
NaOH and 5.7 ml  glacial  acetic  acid to 900 ml  with  organic-free reagent
water.  Dilute to 1 liter with organic-free reagent water.  The pH should
be 4.93 ± 0.02.

5.6  Procedure 2 reagents

     5.6.1   Acetonitrile, CH3CN - HPLC grade.

     5.6.2   2,4-Dinitrophenylhydrazine, C6HaN404,  (DNPH)  - recrystallize
at least twice  with HPLC-grade acetonitrile using procedure in Appendix A.

5.7  Stock standard solutions for Procedure 1

     5.7.1   Stock  formaldehyde  (approximately  1000 mg/L) -  Prepare by
diluting  an   appropriate   amount  of  the  certified   or  standardized
formaldehyde (approximately 265  /zL) to 100 mL  with  organic-free reagent
water.  If a certified formaldehyde solution is not available or there is
any question regarding the quality  of  a  certified  solution,  the solution
may be standardized using the procedure in Sec. 5.7.1.1.

             5.7.1.1   Standardization  of  formaldehyde  stock  solution  -
     Transfer a  25  mL  aliquot of a  0.1  M Na2S03  solution  to  a beaker and
     record the  pH.   Add  a  25.0  mL aliquot  of the formaldehyde stock
     solution (Sec. 5.18.1) and record the pH.   Titrate  this mixture back
     to the original pH  using 0.1 N HC1.   The  formaldehyde concentration
     is calculated  using the following equation:

                 *  *•   ,  /, x    (30.03)(N HCl)(mL HC1)
           Concentration (mg/L)  = -	—	—	L
                                          25.0  mL

     where:

     N HC1  =   Normality of HC1  solution used (in milli-equivalents/mL)
               (1 mmole of HC1 = 1 milli-equivalent of HC1)
     mL HC1 =   mL of standardized HC1 solution used
     30.03 =   Molecular of weight of formaldehyde (in mg/mmole)
                             8315A - 7                         Revision 1
                                                             January 1995

-------
          5.7.2    Stock  aldehyde(s)  and  ketone(s)  -  Prepare by  adding  an
     appropriate amount of the pure material to 90 ml of acetonitrile and dilute
     to 100 ml, to give a  final concentration of  1000 mg/L.

     5.8  Stock standard solutions for Procedure  2

          5.8.1    Preparation of the DNPH  Derivatives for HPLC analysis

                   5.8.1.1   To  a  portion  of  the  recrystallized  DNPH,  add
          sufficient 2N HC1 to obtain  an  approximately saturated solution.  Add
          to  this  solution the target  analyte in molar  excess  of the  DNPH.
          Filter the DNPH  derivative precipitate, wash it with 2N HC1, wash it
          again with water, and allow it to dry in air.

                   5.8.1.2   Check  the  purity  of the DNPH derivative by melting
          point determination or HPLC analysis.   If the impurity  level  is not
          acceptable, recrystallize the derivative in acetonitrile.  Repeat the
          purity check and recrystallization  as  necessary  until  99% purity is
          achieved.

          5.8.2    Preparation of DNPH  derivative  standards and  calibration
     standards for HPLC analysis

                   5.8.2.1   Stock  standard  solutions - Prepare individual stock
          standard solutions  for each  of the target analyte DNPH derivatives by
          dissolving  accurately weighed amounts  in acetonitrile.   Individual
          stock  solutions  of  approximately  100 mg/L may  be  prepared  by
          dissolving 0.010 g of the  solid derivative in 100 mL of acetonitrile.

                   5.8.2.2   Secondary   dilution   standard(s)  -   Using   the
          individual  stock  standard   solutions,  prepare  secondary  dilution
          standards  in  acetonitrile containing the DNPH derivatives  from the
          target  analytes  mixed  together.    Solutions of  100  jug/L may  be
          prepared  by placing  100  /iL  of a  100 mg/L  solution  in  a  100  mL
          volumetric  flask and  diluting  to the mark with acetonitrile.

                   5.8.2.3   Calibration   standards   -   Prepare   a   working
          calibration standard mix from the secondary dilution standard, using
          the mixture of DNPH derivatives  at  concentrations  of 0.5 -  2.0 M9/L
          (which spans the concentration of interest for most indoor air work).
          The concentration of the DNPH derivative in the standard mix solutions
          may need to be adjusted to reflect relative concentration distribution
          in  a real  sample.

     5.9  Standard storage - Store all standard solutions at 4°C in  a glass vial
with a  Teflon®-!ined  cap, leaving minimum headspace, and in the dark.  Standards
should be stable for  about  6 weeks.  All  standards should be checked frequently
for signs of  degradation  or  evaporation,  especially  just prior  to preparing
calibration standards from them.

     5.10 Calibration standards

          Prepare  calibration solutions at a minimum of  5  concentrations for
     each analyte of interest  in organic-free reagent water (or acetonitrile for

                                  8315A -  8                         Revision  1
                                                                  January 1995

-------
      Procedure  2) from the stock standard solution.  The lowest  concentration
      of  each  analyte  should  be at,  or just  above,  the MDLs listed in Tables 1
      or  2.  The other concentrations of the calibration curve should correspond
      to  the expected  range of  concentrations  found  in real  samples.


6.0   SAMPLE COLLECTION,  PRESERVATION,  AND HANDLING

      6.1   See the introductory material to this  Chapter, Organic Analytes, Sec.
4.1.

      6.2   Samples  must  be  refrigerated  at  4°C.    Aqueous  samples must  be
derivatized and extracted within 3 days of sample collection.  The holding times
of leachates  of solid samples should  be kept at a minimum.   All  derivatized
sample extracts should be analyzed within 3  days after preparation.

      6.3   Samples collected  by Methods 0011 or 0100 must  be  refrigerated at
4*C.  It  is recommended that  samples be extracted and analyzed within 30 days of
collection.


7.0   PROCEDURE

      7.1   Extraction  of  solid  samples  (Procedure 1)

           7.1.1  All  solid  samples should  be made as homogeneous as possible
      by  stirring and  removal of  sticks,  rocks,  and other extraneous material.
      When  the sample  is  not  dry, determine the dry weight of the sample, using
      a representative aliquot.  If particle size reduction is necessary, proceed
      as  per Method 1311.

                  7.1.1.1  Determination of  dry weight -  In certain  cases,
           sample results are desired based  on a dry weight basis.   When such
           data  are  desired  or required,  a  portion  of  sample for  dry  weight
           determination  should be weighed out at the same  time as the portion
           used  for analytical  determination.

WARNING;    The drying oven should be contained  in a  hood or vented.  Significant
           laboratory contamination may result from drying a heavily contaminated
           hazardous waste sample.

                  7.1.1.2  Immediately  after   weighing   the   sample   for
           extraction,  weigh  5 -  10 g  of  the  sample  into a  tared  crucible.
           Determine the  % dry  weight  of the sample by drying  overnight  at
           105°C.  Allow  to cool in  a desiccator before weighing:
                      <,/ j     •  Li            sample  ...
                      % dry weight = - - - - 1 — xlOO
                                       g of sample

          7.1.2   Measure  25 g  of solid into a  500-mL  bottle  with a Teflon®-
     lined  screw  cap or crimp top,  and  add 500 mL of  extraction  fluid (Sec.
     5.5.15).  Extract the solid by rotating the  bottle  at approximately 30 rpm
     for     18 hours.  Filter the extract through glass fiber filter paper and
     store  in  sealed bottles at  4°C.   Each mL of extract  represents  0.050 g

                                   8315A  - 9                        Revision 1
                                                                  January 1995

-------
     solid.  Smaller quantities of solid sample may be used with correspondingly
     reduced  volumes  of extraction  fluid maintaining the  1:20 mass  to volume
     ratio.

     7.2  Cleanup  and  separation (Procedure 1)

          7.2.1    Cleanup procedures may not be necessary for a relatively clean
     sample matrix.  The cleanup procedures recommended in this method have been
     used  for the analysis  of  various sample  types.    If  particular  samples
     demand  the use  of an  alternative cleanup  procedure,  the  analyst  must
     determine  the  elution  profile  and  demonstrate  that the  recovery  of
     formaldehyde  from a spiked  sample is greater than  85%.   Recovery may be
     lower for  samples  which  form emulsions.

          7.2.2    If the sample is not clear,  or the complexity is unknown, the
     entire sample should be centrifuged at 2500 rpm for 10  minutes.  Decant the
     supernatant liquid from the centrifuge bottle, and filter  through glass
     fiber filter  paper into  a container which  can be tightly sealed.

     7.3  Derivatization  (Procedure 1)

          7.3.1    For  aqueous samples,  measure an aliquot  of  sample which is
     appropriate  to the  anticipated  analyte  concentration range  (nominally
     100 mL).  Quantitatively transfer the sample aliquot to the reaction vessel
     (Sec. 4.2).

          7.3.2    For  solid  samples,   1  to  10  ml  of  extract (Sec.  7.1)  will
     usually  be required.   The  amount used  for  a particular sample  must be
     determined through preliminary experiments.

NOTE: In cases where the selected sample or extract volume  is less than 100 ml,
      the total volume  of the aqueous layer should  be  adjusted  to 100  ml with
      organic-free  reagent  water.     Record  original  sample  volume prior to
      dilution.

          7.3.3    Derivatization and  extraction of the target  analytes may be
     accomplished  using the liquid-solid (Sec. 7.3.4)  or  liquid-liquid (Sec.
     7.3.5) procedures.

          7.3.4    Liquid-solid derivatization  and  extraction

                   7.3.4.1   For  analytes other  than formaldehyde,  add  4 ml of
          citrate  buffer and  adjust the pH to 3.0  ± 0.1  with 6M HC1 or 6M NaOH.
          Add 6 mL of  DNPH  reagent,  seal  the  container, and place in a heated
           (40eC),  orbital shaker (Sec. 4.2.12)  for 1 hour.  Adjust the agitation
          to  produce a gentle swirling  of the  reaction  solution.

                   7.3.4.2   If formaldehyde is the only analyte of interest, add
          4 mL acetate buffer and adjust pH to  5.0 ± 0.1 with 6M HC1 or 6M NaOH.
          Add 6 mL of  DNPH  reagent,  seal  the  container, and place in a heated
           (40eC),  orbital shaker (Sec. 4.2.12)  for 1 hour.  Adjust the agitation
          to  produce a gentle swirling  of the  reaction  solution.
                                  8315A - 10                        Revision 1
                                                                  January 1995

-------
                   7.3.4.3   Assemble the vacuum manifold and connect to a water
           aspirator  or vacuum pump.  Attach  a  2-g sorbent cartridge   to the
           vacuum  manifold.   Condition each cartridge by passing  10  mL dilute
           citrate  buffer  (10 ml of 1 M citrate buffer dissolved  in  250 mL of
           organic-free  reagent  water) through each sorbent cartridge.

                   7.3.4.4   Remove  the   reaction   vessel   from   the   shaker
           immediately  at the  end of the one hour reaction period and add 10 ml
           saturated  NaCl solution  to  the  vessel.

                   7.3.4.5   Quantitatively transfer the reaction solution to the
           sorbent  cartridge and apply a vacuum so that the  solution  is drawn
           through  the cartridge at a rate of 3 to 5 mL/min.  Continue applying
           the  vacuum for about  1  minute  after the  liquid sample has  passed
           through  the cartridge.

                   7.3.4.6   While maintaining the vacuum conditions described in
           Sec.  7.3.4.4, elute each cartridge  train with  approximately  9 mL of
           acetonitrile   directly into  a  10 mL  volumetric flask.  Dilute the
           solution to volume  with  acetonitrile,  mix thoroughly, and place in a
           tightly  sealed vial until analyzed.

NOTE: Because this method uses an excess of DNPH,  the cartridges will  remain a
      yellow color after completion of Sec. 7.3.4.5.   The presence of this color
      is not indicative of the loss of the analyte derivatives.

           7.3.5    Liquid-liquid derivatization and extraction

                   7.3.5.1   For  analytes other than formaldehyde,  add  4  mL of
           citrate  buffer and  adjust the pH to  3.0  ± 0.1 with 6M HC1 or 6M NaOH.
           Add  6 mL of  DNPH  reagent,  seal  the  container,  and  place in  a heated
           (40°C),  orbital shaker  for  1  hour.  Adjust  the  agitation to  produce
           a gentle swirling  of  the reaction solution.

                   7.3.5.2   If formaldehyde is the only analyte of interest, add
           4 mL acetate buffer and  adjust pH to 5.0 ± 0.1 with 6M HC1 or 6M NaOH.
           Add  6 mL of  DNPH  reagent,  seal  the  container,  and  place in  a heated
           (40°C),  orbital shaker  for  1  hour.  Adjust  the  agitation to  produce
           a gentle swirling  of  the reaction solution.

                   7.3.5.3   Serially extract  the  solution  with  three  20  mL
           portions of methylene chloride using a 250 mL separatory funnel.  If
           an  emulsion  forms  upon  extraction, remove  the entire  emulsion and
           centrifuge  at 2000 rpm  for  10  minutes.    Separate  the layers  and
           proceed  with  the  next extraction.   Combine the methylene  chloride
           layers in a 125-mL Erlenmeyer flask containing 5.0 grams of anhydrous
           sodium  sulfate.     Swirl  contents  to  complete  the  extract  drying
           process.

                   7.3.5.4   Assemble  a  Kuderna-Danish (K-D)  concentrator  by
           attaching a 10-mL  concentrator tube to a  500-mL evaporator flask.  See
           Sec. 4.0 of Method  3510.  Pour the extract  into the evaporator flask
           being careful to minimize transfer of sodium sulfate granules.   Wash


                                  8315A -  11                         Revision 1
                                                                  January 1995

-------
          the Erlenmeyer flask with 30 ml of methylene chloride and add wash to
          the evaporator flask to complete quantitative transfer.

                   7.3.5.5   Concentrate the  extract to  a final  volume of 5 ml,
          using the K-D techniques, as described in Method 3510.   Exchange the
          solvent  to acetonitrile prior  to analysis.

     7.4  Extraction of samples from Methods 0011 and 0100 (Procedures 1 and 2)

          7.4.1    Stack gas  samples collected by Method 0011 (Procedure  1)

                   7.4.1.1   Measure  the  volume  of  the  aqueous  phase of  the
          sample prior to extraction (for moisture determination when the volume
          was not  measured  in the field).  Pour the sample  into  a separatory
          funnel and drain the methylene  chloride  into a volumetric flask.

                   7.4.1.2   Extract  the  aqueous solution  with  two  or  three
          aliquots of methylene chloride.  Add the methylene chloride extracts
          to the volumetric  flask.

                   7.4.1.3   Fill the volumetric flask to  the line with methylene
          chloride.  Mix well and remove  an aliquot.

                   7.4.1.4   If high concentrations  of formaldehyde are present,
          the extract can be diluted  with mobile phase,  otherwise the extract
          solvent  must  be  exchanged  as  described in  Sec.  7.3.5.5.    If  low
          concentrations  of  formaldehyde are  present,  the  sample  should  be
          concentrated during the solvent exchange procedure.

                   7.4.1.5   Store  the sample at  4°C.   If the extract  will  be
          stored longer than  two days, it should be transferred to a vial with
          a Teflon®-! ined screw cap, or a crimp top  with a Teflon®-! ined septum.
          Proceed  with HPLC  chromatographic analysis if further cleanup is not
          required.

          7.4.2    Ambient air samples  collected by Method 0100 (Procedure 2)

                   7.4.2.1   The samples will be received by the laboratory in a
          friction-top can containing 2 - 5 cm of granular charcoal, and should
          be  stored   in   this  can,   in  a  refrigerator,  until   analysis.
          Alternatively,  the  samples  may  also   be  stored  alone  in  their
          individual glass containers.  The time between sampling and analysis
          should not exceed  30 days.

                   7.4.2.2   Remove the sample cartridge from the labeled culture
          tube.    Connect  the  sample  cartridge  (outlet  or long end  during
          sampling) to a clean syringe.

NOTE: The liquid flow during  desorption should  be in the opposite direction from
      the air flow during sample collection  (i.e,  backflush the cartridge).

                   7.4.2.3   Place  the  cartridge/syringe  in the  syringe rack.
                                  8315A - 12                        Revision 1
                                                                  January 1995

-------
                   7.4.2.4   Backflush the cartridge  (gravity  feed)  by passing
           6 ml  of acetonitrile  from the syringe  through  the cartridge  to a
           graduated  test tube, or  to a  5-mL volumetric  flask.

NOTE: A dry cartridge has an acetonitrile holdup volume  slightly greater than 1
      ml.  The eluate flow  may stop  before  the  acetonitrile  in the syringe is
      completely drained into the cartridge because of air trapped between the
      cartridge filter and  the syringe Luer-Lok tip.   If this happens, displace
      the  trapped  air  with the  acetonitrile  in  the  syringe  using  a long-tip
      disposable Pasteur pipet.

                   7.4.2.5   Dilute to the 5 ml mark  with  acetonitrile.   Label
           the flask with sample identification.  Pipet two aliquots into sample
           vials having Teflon®-!ined septa.

                   7.4.2.6   Store  the  sample at  4°C.    Proceed  with  HPLC
           chromatographic  analysis of the first  aliquot if further cleanup is
           not required.  Store the second aliquot in  the refrigerator until the
           results of the first aliquot analysis are complete and validated.  The
           second aliquot can  be  used for confirmatory analysis, if necessary.

     7.5   Recommended chromatographic conditions

           7.5.1    Procedure 1 - For aqueous samples,  soil or waste samples, and
     stack gas samples collected by  Method 0011.
     Column:

     Mobile Phase Gradient:
     Flow Rate:

     Detector:

     Injection Volume:
CIS, 4.6 mm x 250 mm ID, 5 jum particle size

70/30 acetonitrile/water (v/v),  hold for 20 min.
70/30 acetonitrile/water to  100% acetonitrile in
15 min.
100% acetonitrile for 15 min.

1.2 mL/min

Ultraviolet, operated at 360 nm

20 ML
          7.5.2    Procedure  2 - For  ambient  air samples collected  by Method
     0100.

     Column:


     Mobile Phase Gradient:
Two HPLC columns,  4.6  mm x 250  mm  ID,  (Zorbax
CDS, or equivalent) in series

60/40 CH3CN/H20,  hold for 0 min.
60/40 to 75/25 CH3CN/H20,  linearly in  30 min.
75/25 to 100% CH3CN,  linearly in  20  min.
100% CH3CN  for 5  minutes.
100% to 60/40 CH3CN/H20,  linearly in 1 min.
60/40 CH3CN/H20 for 15  minutes.
     Detector:
Ultraviolet, operated at 360 nm
                                  8315A - 13
                                     Revision 1
                                   January 1995

-------
     Flow Rate:                1.0 mL/min

     Sample  Injection volume:  25 fj.1 (suggested)

NOTE: Analysts  are  advised   to  adjust  their  HPLC   systems   to  optimize
      chromatographic conditions  for  their particular analytical needs.   The
      separation of acrolein,  acetone,  and propionaldehyde should be a minimum
      criterion of the  optimization in Procedure 2.

          7.5.3    Filter and degas the mobile phase to remove dissolved gasses,
     using the following procedure:

                   7.5.3.1   Filter each solvent (water and acetonitrile) through
          a  0.22  jum polyester membrane  filter, in  an  all glass and Teflon®
          suction  filtration apparatus.

                   7.5.3.2   Degas each  filtered  solution  by purging with helium
          for  10  - 15  minutes  (100 mL/min)  or  by  heating to 60°C  for  5  -  10
          minutes  in an Erlenmeyer flask  covered with a watch glass.  A constant
          back pressure restrictor (350 kPa)  or  15 - 30 cm of 0.25 mm ID Teflon®
          tubing should be placed after the detector to eliminate further mobile
          phase outgassing.

                   7.5.3.3   Place  the   mobile   phase   components   in   their
          respective HPLC  solvent reservoirs,  and  program the gradient system
          according to  the conditions  listed in Sec. 7.5.2.   Allow the system
          to pump  for  20  - 30  minutes at  a  flow rate of  1.0 mL/min  with the
          initial  solvent  mixture ratio  (60%/40%  CH3CN/H20).    Display  the
          detector output  on a strip chart recorder or similar output device to
          establish a stable baseline.

     7.6  Calibration

          7.6.1    Establish liquid  chromatographic   operating  conditions  to
     produce a retention  time  similar to that  indicated  in  Table 1  for the
     liquid-solid  derivatization and extraction or  in Table 2 for liquid-liquid
     derivatization  and  extraction.     For  determination of retention  time
     windows, see Sec.  7.5  of Method 8000.  Suggested  chromatographic conditions
     are provided  in Sec.  7.5.

          7.6.2    Process   each  calibration   standard   solution   through
     derivatization and extraction, using the same procedure employed for sample
     processing (Sees.  7.3.4 or 7.3.5).

          7.6.3    Analyze  a solvent blank to  ensure that the system is  clean and
     interference  free.

NOTE: The samples  and standards must be  allowed to come  to ambient temperature
      before analysis.

          7.6.4    Analyze   each  processed  calibration   standard  using  the
     chromatographic  conditions  listed  in Sec. 7.5,  and tabulate  peak  area
     against calibration solution concentration  in
                                  8315A - 14                        Revision 1
                                                                  January 1995

-------
     7.6.5   Tabulate  the peak  area along with  standard  concentration
injected to determine the calibration factor (CF)  for the  analyte at  each
concentration  (see Sec.  7.8.1  for  equations).   The  percent  relative
standard deviation  (%RSD) of  the  mean  CF  of  the calibration  standards
should be < 20 percent or a system check will  have to be performed.   If a
calibration check  after the  system check does  not  meet the criteria,  a
recalibration will have  to be  performed.   If the recalibration  does  not
meet the established criteria,  new calibration standards must be made.

     7.6.6   The  working calibration curve  must  be  verified  each  day,
before  and after analyses  are  performed,  by  analyzing  one  or  more
calibration standards.  The calibration factor obtained should fall  within
± 15 percent of the initially  established calibration factor or  a  system
check will  have to be performed.   If a calibration check after  the  system
check does not meet the criteria, the system must  be recalibrated.

     7.6.7   After  10  sample  runs,  or  less,  one  of  the calibration
standards must be  reanalyzed to ensure that the DNPH derivative calibration
factors remain within ± 15% of the original  calibration factors.

7.7  Sample analysis

     7.7.1   Analyze samples  by HPLC,  using conditions established in  Sec.
7.5.  For Procedure 1 analytes,  Tables 1  and  2 list the retention times and
MDLs that were obtained under these conditions.  For Procedure 2 analytes,
refer to Figure 3 for the sample chromatogram.

     7.7.2   If the peak area exceeds the linear range of  the calibration
curve, a smaller  sample  injection volume should be used.  Alternatively,
the final solution may be diluted with acetonitrile and reanalyzed.

     7.7.3   After  elution  of   the  target  analytes,  calculate   the
concentration of analytes found  in the samples  using  the  equations found
in Sec. 7.8 or the specific sampling method used.

     7.7.4   If the peak area measurement is prevented by  the presence of
observed interferences, further cleanup  is required.

7.8  Calculations

     7.8.1   Calculate each calibration factor, mean  calibration factor,
standard deviation, and percent  relative  standard deviation as follows:


       _  _     Peak Area of  the  Compound in the Standard
             Concentration of the Compound Injected (in ug/L)
                             8315A  -  15                        Revision  1
                                                            January  1995

-------
                            mean CF  = CF =
SD =
       V(CF.-CF)2
       ^    '                                SD
RSD = — x 100
           n-1                               CF
    where:
    CF   *  Mean calibration factor using the 5 calibration concentrations.
    CFj   =  Calibration factor for calibration standard i (i = 1-5).
    RSD  =  Relative standard deviation of the calibration factors.
    n    =  Number of calibration standards.
         7.8.2   Calculate the concentrations  in liquid samples as  follows:
      Concentration of aldehydes (ug/L) = (Area of sample peak) x 100
                                                    IT x Vs
    where:
         CF   =  Mean calibration factor the analyte.
         Vs   =  Number of ml of sample (unitless).
         7.8.3   Calculate the concentration in solid samples  as follows:
      Concentration of aldehydes (ug/g) = (Area of sample peak) x 100
    where:
         CF   =  Mean calibration factor the analyte.
         Vex   =  Number of ml extraction fluid aliquot (unitless).
         7.8.4   Calculate  the  concentration  of formaldehyde  in stack  gas
    samples (Method 0011) as follows: (Procedure 1)
                 7.8.4.1  Calculation of total formaldehyde
                 To determine the total  formaldehyde in mg,  use the following
         equation:
 -r .  ,    f    -,...    ,.    „   nr     (g/mole formaldehyde)      ...3    ,
 Total mg formaldehyde = C. x V x DF  x——	1—'.— x 10 d mg/ug
                         d           (g/mole DNPH derivatiibe)
                                 8315A -  16                        Revision 1
                                                                 January 1995

-------
where:
Cd = measured concentration of DNPH-formaldehyde  derivative,  mg/L
V  = organic extract volume, ml
DF = dilution factor

             7.8.4.2   Formaldehyde concentration in stack gas

             Determine  the formaldehyde  concentration in  the  stack gas
     using the following equation:
                 _   K x (total formaldehyde in mg)
                                   m(std)
     where:

     K

     "mlstd)
= 35.31  ft3/")3, if Vm(std) is expressed in English units
= 1.00 m3/m3, if Vm(8td) is expressed in metric units
= volume of  gas  sample  as  measured  by  dry  gas  meter,
  corrected  to  standard conditions,  dscm (dscf)
     7.8.5   Calculation  of the concentration  of  formaldehyde and other
carbonyls from indoor air sampling by Method 0100.  (Procedure 2)

             7.8.5.1   The concentration of target analyte "a," in air at
     standard  conditions  (25°C  and  101.3 kPa),  Conc8td  in  ng/L,  may be
     calculated using the following equation:
                .n      _  (Areaa)(Vo1a)(MWa)(1000 ng/ug) ^ ^
                            (RF)(MWd)(VTotStd )(1000inl/L)
     where:

     Areaa
     CF

     Vol.
     MWa
     MWd

     "TotStd
     DF
                                             "a"
from  the
                                                          in
Area of the sample peak for analyte  "a"
Mean  calibration   factor  for  analyte
calibration in M9/L.  (See Sec.  7.8.1)
Total volume of the sample cartridge eluate  (ml)
Molecular weight of analyte "a" in g/mole
Molecular weight of the DNPH derivative of analyte
g/mole
Total  volume  of   air   sampled  converted  to  standard
conditions in liters (L).  (To calculate the concentration
at sampling conditions use Vtot.) (See Sec. 9.1.3 of Method
0100)
Dilution Factor for the  sample  cartridge eluate,  if any.
If there is no dilution, DF = 1
                             8315A - 17
                                                  Revision  1
                                               January  1995

-------
                   7.8.5.2   The target  analyte  "a" concentration  at  standard
          conditions may be  converted to parts per billion by volume, Conca in
          ppbv,  using the  following  equation:
                       Conca  ,in ppbv =
                                        (Conco_,td)(22.4)
                                             (MM.)


          where:

          C°nca.std  = Concentration of  "a," at standard conditions,  in ng/L
          22.4     = Ideal  gas  law  volume  (22.4 nl_  of  gas  =  1  nmole,  at
                     standard  conditions)
          MWa      = Molecular weight  of analyte  "a" in g/mole (or  ng/nmole)
8.0  QUALITY CONTROL

     8.1  Refer to Chapter One and Method 8000 for specific quality control (QC)
procedures.  Each  laboratory should maintain a formal quality assurance program.
The laboratory should also maintain  records to document the quality of the data
generated.

     8.2   Quality  control  procedures necessary  to  evaluate the HPLC  system
operation are found in Method 8000, Sec. 7.0 and  include evaluation of retention
time windows, calibration verification and  chromatographic analysis of samples.

     8.3     Initial  Demonstration  of  Proficiency  -  Each  laboratory  must
demonstrate initial proficiency with each sample preparation and determinative
method combination  it utilizes, by  generating data  of acceptable accuracy and
precision for target analytes in a clean matrix.  The laboratory must also repeat
the following operations whenever new staff are trained or significant changes
in instrumentation are made.  See Method 8000, Sec.  8.0 for information on how
to accomplish this demonstration.

     8.4   Sample  Quality  Control for  Preparation and Analysis - The laboratory
must also  have  procedures for documenting the effect of the  matrix  on method
performance  (precision,  accuracy,  and detection  limit).   At a  minimum,  this
includes the analysis of QC samples  including  a method blank,  a matrix spike, a
duplicate, and a laboratory control  sample (LCS) in each analytical  batch.

           8.4.1    Documenting  the  effect of  the matrix  should include the
     analysis of  at least one matrix spike and one duplicate unspiked sample or
     one  matrix spike/matrix spike  duplicate pair.  The decision on whether to
     prepare  and  analyze duplicate  samples  or  a  matrix  spike/matrix  spike
     duplicate must be based on a knowledge of the samples in  the sample batch.
     If samples are expected to contain target analytes, then laboratories may
     use  one matrix spike and a duplicate analysis of  an unspiked field sample.
     If samples are not  expected to contain target analytes, laboratories should
     use  a matrix spike  and matrix  spike duplicate pair.
                                  8315A - 18                        Revision 1
                                                                  January 1995

-------
           8.4.2  A Laboratory Control Sample (LCS)  should be included with each
      analytical  batch.   The LCS  consists  of an aliquot of a  clean (control)
      matrix  similar to the sample matrix and of the same weight or volume.  The
      LCS  is  spiked with the same analytes at the  same concentrations  as the
      matrix  spike.   When the results of the  matrix spike  analysis  indicate a
      potential problem due to the  sample matrix itself, the  LCS results are used
      to verify that the  laboratory can perform the  analysis in a clean matrix.

           8,4.3    Refer  to Table 4 for QC  acceptance  limits  derived from the
      interlaboratory method  validation  study  on Method 8315.

           8.4.4   See Method 8000,  Sec. 8.0  for  the details on  carrying out
      sample  quality control  procedures  for  preparation and analysis.

      8.5    It  is  recommended  that  the laboratory adopt  additional  quality
assurance practices for use with this method.  The specific practices that are
most productive depend upon the needs of the  laboratory and the  nature  of the
samples.  Whenever possible, the  laboratory should analyze standard reference
materials and participate in relevant performance evaluation studies.


9.0   METHOD  PERFORMANCE

      9.1   The  MDLs for  Procedure 1  listed  in  Table 1  were obtained  using
organic-free reagent water and  liquid-solid extraction.  The MDLs for Procedure
1 listed in Table 2 were obtained using organic-free reagent water and methylene
chloride extraction.  Results  reported  in  Tables  1  and 2  were  achieved using
fortified  reagent  water volumes  of  100 mL.   Lower  detection  limits  may  be
obtained using larger sample volumes.

           9.1.1    Procedure  1  of  this  method has  been tested  for linearity of
      recovery from spiked organic-free reagent water and has been demonstrated
      to be applicable over the range 50-1000  jug/L  .

           9.1.2    To generate  the MDL and precision and accuracy data reported
      in this section, analytes were  segregated into two  spiking  groups,  A and
      B.   Chromatograms  using  liquid-solid  and  liquid-liquid extraction are
      presented in  Figures  1  (a and b) and  2  (a and b), respectively.

      9.2   The sensitivity of Procedure 2 sampling  (Method 0100) and analysis is
listed in Table 3.

      9.3   Method 8315, Procedure 1,  was tested by 12 laboratories using reagent
water and  ground waters  spiked at six  concentrations over the  range  30-2200
p.g/1.  Method accuracy and precision were  found  to be directly  related  to the
concentration of the analyte and independent of the sample matrix.  Mean recovery
weighted  linear  regression  equations,  calculated  as a  function  of  spike
concentration,  as  well  as  overall  and  single-analyst  precision  regression
equations, calculated as functions of mean recovery, are  presented in Table 5.
These equations  can  be  used to  estimate mean  recovery  and precision  at any
concentration value within the  range tested.
                                  8315A - 19                        Revision 1
                                                                  January 1995

-------
10.0 REFERENCES

1.   "OSHA  Safety and Health Standards,  General  Industry",  (29CRF1910).
     Occupational Safety and Health Administration, OSHA 2206, (Revised, January
     1976).


11.0 SAFETY

     11.1 The  toxicity  or carcinogenicity of each reagent used in this method
has not been precisely defined; however, each chemical compound should be treated
as a potential  health hazard.  From this viewpoint, exposure to these chemicals
must be reduced to the lowest possible level by whatever means  available.  The
laboratory  is  responsible  for  maintaining a current  awareness file  of OSHA
regulations regarding  the  safe  handling  of  the  chemicals specified  in this
method.  A  reference file of material safety data sheets  should  also be made
available to  all  personnel involved  in  the chemical  analysis.    Additional
references to laboratory safety are available.

     11.2 Formaldehyde has  been tentatively classified as  a known or suspected,
human or mammalian carcinogen.
                                  8315A - 20                        Revision 1
                                                                  January 1995

-------
                                    TABLE 1

                 PROCEDURE 1  - METHOD DETECTION LIMITS" USING
                            LIQUID-SOLID EXTRACTION
                       Retention Time                MDL
Analyte                   (minutes)                 (/ig/L)"
Formaldehyde                 5.3                     6.2
Acetaldehyde                 7.4                    43.7b
Propanal                    11.7                    11.0
Crotonaldehyde              16.1                     5.9
Butanal                     18.1                     6.3
Cyclohexanone               27.6                     5.8
Pentanal                    28.4                    15.3
Hexanal                     34.1                    10.7
Heptanal                    35.0                    10.0
Octanal                     40.1                     6.9
Nonanal                     40.4                    13.6
Decanal                     44.1                     4.4
" The method detection limit (MDL) is defined in Chapter One.  With the exception
  of acetaldehyde, all reported MDLs are based upon analyses of 6 to 8 replicate
  blanks spiked at 25 M9/L-

b The reported MDL is based upon analyses of three replicate blanks fortified at
  250 M9/L.
                                  8315A - 21                        Revision 1
                                                                  January 1995

-------
                                    TABLE  2

                 PROCEDURE 1 - METHOD DETECTION LIMITS8 USING
                           LIQUID-LIQUID EXTRACTION
                       Retention Time                MDL
Analyte                   (minutes)
Formaldehyde
Acetaldehyde
Propanal
Crotonaldehyde
Butanal
Cyclohexanone
Pentanal
Hexanal
Heptanal
Octanal
Nonanal
Decanal
5.3
7.4
11.7
16.1
18.1
27.6
28.4
34.1
35.0
40.1
40.4
44.1
23.2
110.2"
8.4
5.9
7.8
6.9
13.4
12.4
6.6
9.9
7.4
13.1
  The method detection limit (MDL) is defined in Chapter One.  With the exception
  of acetaldehyde, all reported MDLs  are based upon analyses of 6 to 8 replicate
  blanks spiked at 25
b The reported MDL is  based upon analyses of three replicate blanks fortified at
  250
                                  8315A  - 22                        Revision 1
                                                                  January  1995

-------
                                    TABLE  3

       PROCEDURE 2  -  SENSITIVITY (ppb,  v/v)  OF  SAMPLING AND ANALYSIS  FOR
        CARBONYL COMPOUNDS IN AMBIENT AIR USING AN ADSORBENT CARTRIDGE
                          FOLLOWED BY GRADIENT HPLC"
Compound
10    20
30
 Sample  Volume  (L)b

40    50    100   200
300   400   500
Acetaldehyde
Acetone
Acrolein
Benzaldehyde
Butyraldehyde
Crotonaldehyde
2,5-Dimethyl-
benzaldehyde
Formaldehyde
Hexanal
Isovaleraldehyde
Propionaldehyde
m-Tolualdehyde
o-Tolualdehyde
p-Tolualdehyde
Valeraldehyde
1.36
1.28
1.29
1.07
1.21
1.22

0.97
1.45
1.09
1.15
1.28
1.02
1.02
1.02
1.15
0.68
0.64
0.65
0.53
0.61
0.61

0.49
0.73
0.55
0.57
0.64
0.51
0.51
0.51
0.57
0.45
0.43
0.43
0.36
0.40
0.41

0.32
0.48
0.36
0.38
0.43
0.34
0.34
0.34
0.38
0.34
0.32
0.32
0.27
0.30
0.31

0.24
0.36
0.27
0.29
0.32
0.25
0.25
0.25
0.29
0.27
0.26
0.26
0.21
0.24
0.24

0.19
0.29
0.22
0.23
0.26
0.20
0.20
0.20
0.23
0.
0.
0.
0.
0.
0.

0.
0.
0.
0.
0.
0.
0.
0.
0.
14
13
13
11
12
12

10
15
11
11
13
10
10
10
11
0.07
0.06
0.06
0.05
0.06
0.06

0.05
0.07
0.05
0.06
0.06
0.05
0.05
0.05
0.06
0.05
0.04
0.04
0.04
0.04
0.04

0.03
0.05
0.04
0.04
0.04
0.03
0.03
0.03
0.04
0.03
0.03
0.03
0.03
0.03
0.03

0.02
0.04
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.02
0.02
0.02

0.02
0.03
0.02
0.02
0.03
0.02
0.02
0.02
0.02
8 The ppb values are measured at 1  atm  and 25°C. The sample cartridge is eluted
  with 5  mL acetonitrile and  25 juL is  injected  into the HPLC.   The  maximum
  sampling flow through a DNPH-coated Sep-PAK is about 1.5 L/minute.
b A sample volume of 1000 L was also analyzed.
  of 0.01 ppb for all the target analytes.
                             The results show a sensitivity
                                  8315A - 23
                                                 Revision  1
                                               January 1995

-------
                                   TABLE 4

               PERFORMANCE-BASED QC ACCEPTANCE  LIMITS CALCULATED
                      USING THE COLLABORATIVE STUDY DATA
Spike
Analyte Concentration* Xb SRC
Formaldehyde
Propanal
Crotonaldehyde
Butanal
Cyclohexanone
Hexanal
Octanal
Decanal
160
160
160
160
160
160
160
160
154
148
160
151
169
151
145
153
30.5
22.4
34.8
22.7
39.2
34.6
40.1
40.0
Acceptance
Limits, %d
39
50
35
52
32
30
15
21
-153
-134
-165
-137
-179
-159
-166
-171
" Spike concentration, /xg/L.

b Mean  recovery calculated  using the  reagent water,  mean  recovery,  linear
  regression equation, M9/L.

0 Overall  standard  deviation calculated  using  the  reagent water,  overall
  standard deviation linear regression equation, p.g/1.

d Acceptance limits calculated as (X ± 3SR)100/spike concentration.
                                  8315A  - 24
  Revision 1
January 1995

-------
                                   TABLE 5

 WEIGHTED  LINEAR REGRESSION EQUATIONS FOR MEAN RECOVERY AND PRECISION (/ig/L)
Applicable
Analyte Cone. Range
Formaldehyde 39.2 - 2450
Propanal 31.9 - 2000
Crotonaldehyde 32.4 - 2030
Butanal 35.4 - 2220
Cyclohexanone 31.6 - 1970
Hexanal 34.1- 2130
Octanal 32.9 - 2050
Decanal 33.2 - 2080

X
SR
sr
X
SR
sr
X
SR
X
SR
X
SR
sr
X
SR
Sr
X
SR
X
SR
Reagent Water
0.909C + 8.79
0.185X + 1.98"
0.093X + 5.79
0.858C + 10.49
0.140X + 1.63
0.056X + 2.76
0.975C + 4.36
0.185X + 5.15
0.096X + 1.85
0.902C + 6.65
0.149X + 0.21
0.086X - 0.71
0.962C + 14.97
0.204X + 4.73"
0.187X + 3.46
0.844C + 15.81
0.169X + 9.07
0.098X + 0.37"
0.856C + 7.88
0.200X + 11.17
0.092X + 1.71"
0.883C + 12.00
0.225X + 5.52
0.088X + 2.28"
Ground Water
0.870C + 14.84
0.177X + 13.85
0.108X + 6.24
0.892C + 22.22
0.180X + 12.37
0.146X + 2.08"
0.971C + 2.94
0.157X + 6.09
0.119X - 2.27
0.925C + 12.71
0.140X + 6.89
0.108X - 1.63"
0.946C + 28.95
0.345X + 5.02
0.123X + 7.64
0.926C + 9.16
0.132X + 8.31
0.074X - 0.40"
0.914C + 13.09
0.097X + 12.41
0.039X + 1.14
0.908C + 6.46
0.153X + 2.23
0.052X + 0.37
8  Variance is not constant over concentration  range.
X  Mean recovery,  M9/L, exclusive of outliers.
SR Overall  standard deviation,  M9/U exclusive  of outliers.
sr  Single-analyst  standard deviation,  M9/U  exclusive  of outliers.
                                  8315A -  25                        Revision  1
                                                                  January  1995

-------
                                      FIGURE  la
                                     PROCEDURE 2
          LIQUID-SOLID PROCEDURAL STANDARD OF GROUP A ANALYTES AT 625
  -0.80-
  -1.00-
 ,-1.20-
 o
v -». w
o       <
  -1.80-
  -i.ao-
  -a.oo-
                i.oo
a.oo             3.00
       x  SO1  MinutM
4.00
            Retention  Time  (min)

                    5.33
                   11.68
                   18.13
                   27.93
                   36.60
                   42.99
   Analvte Derivative

        Formaldehyde
        Propanal
        Butanal
        Cyclohexanone
        Heptanal
        Nonanal
                                     8315A - 26
                                      Revision 1
                                    January 1995

-------
                                     FIGURE  Ib
                                    PROCEDURE 1
         LIQUID-SOLID PROCEDURAL STANDARD OF GROUP B ANALYTES  AT  625
 -0.60-
 -0.80-
 -1.00-
2-1.20H
 -1.40-
 -1.60-
 -1.80-
             1.00
a.oo              3.00
         »  101  Minutes
4.00
           Retention Time (min)

                    7.50
                   16.68
                   26.88
                   32.53
                   40.36
                   45.49
     Analvte Derivative

           Acetaldehyde
           Crotonaldehyde
           Pentanal
           Hexanal
           Octanal
           Decanal
                                     8315A  -  27
                                        Revision 1
                                      January 1995

-------
                                     FIGURE 2a

                                    PROCEDURE 1
         LIQUID-LIQUID  PROCEDURAL  STANDARD  OF GROUP  A  ANALYTES  AT  625  /xg/L
-i.40
             1.00
2.00              3.00
        x iO1 ainutts
                                                                  4.00
           Retention Time (min)

                   5.82
                  13.23
                  20.83
                  29.95
                  37.77
                  43.80
     Analvte Derivative

           Formaldehyde
           Propanal
           Butanal
           Cyclohexanone
           Heptanal
           Nonanal
                                    8315A  - 28
                                        Revision 1
                                      January 1995

-------
                                     FIGURE 2b

                                    PROCEDURE  1
         LIQUID-LIQUID  PROCEDURAL  STANDARD  OF  GROUP B ANALYTES AT 625  /jg/L
-8.00H
            1.00
1.00
       3.00
x 10*  •inut««
4.00
          Retention Time (min)

                   7.79
                  17.38
                  27.22
                  32.76
                  40.51
                  45.62
      Analyte Derivative

            Acetaldehyde
            Crotonaldehyde
            Pentanal
            Hexanal
            Octanal
            Decanal
                                    8315A  - 29
                                         Revision 1
                                       January 1995

-------
                                 FIGURE  3
                               PROCEDURE 2
CHROMATOGRAPHIC  SEPARATION OF THE DNPH DERIVATIVES OF 15  CARBONYL COMPOUNDS
            DNPM
                              uuu
                10
                              30
                                            30
                                                          40
                                   TIME. min
                           Peak Identification
        Number

          1
          2
          3
          4
          5
          6
          7
          8
          9
         10
         11
         12
         13
         14
         15
Compound
Concentration  (nq/uL)
Formaldehyde                 1.140
Acetaldehyde                 1.000
Acrolein                     1.000
Acetone                      1.000
Propanal                     1.000
Crotonaldehyde               1.000
Butanal                      0.905
Benzaldehyde                 1.000
Isovaleraldehyde              0.450
Pentanal                     0.485
o-Tolualdehyde               0.515
m-Tolualdehyde               0.505
p-Tolualdehyde               0.510
Hexanal                      1.000
2,4-Dimethylbenzaldehyde      0.510
                               8315A - 30
                                             Revision  1
                                           January  1995

-------
                                     METHOD  8315A
                      DETERMINATION  OF  CARBONYL COMPOUNDS
              BY HIGH  PERFORMANCE LIQUID CHROMATOGRAPHY  (HPLC)
                                          Media (Optionl)
  7.1.1 -7.1.1.1
Homogenize sample
 and determne dry
     weight
  7.1.2 Extract
 sample for 18
 hours; fllter and
  store extract
7 3.2 Measure 1-10
 ml extract; adjust
 volume to 100 mL
   with water
                                      7.0 What is
                                      the sample
                                       matrix?
                                     7.0 Is media
                                       solid or
                                      aqueous''
                                        Is sample
                                     dear or sample
                                       complexity
                                        Known?
                                                     Stack Gas (Option
            0
No
7.2.2 Centrifuge sample
  at 2500 rpm for 10
         ; decant
      and filter
                                                         Aqueous
                7.3.1 Measure
               aliquot of sample;
                adjust volume to
               100 ml with water
                                    7.3.5.5 Exchange
                                   solvent to methanol
                                      0
                                     8315A -  31
                                                                                   Revision 1
                                                                                January  1995

-------
           METHOD 83ISA
             continued
                                                                                 i
1

7.4.2.2 - 7.4.2.3
Connect sample cartridge
to dean syringe and
place in syringe rack
i

7.4.2.4 Backflush
cartridge with
acetonitrite
     7.4.2.4
   Doeseluaie
   flow become
    blocked?
   7.4.24 Displace
    trapped air with
    acetonitrile in
 syringe using a long-tip
disposable Pasteur pipet
 7 4.2.5 Dilute to 5
mL with acetonitrile;
 label flask; pipet 2
   aliquots into
   sample vials
   7.4.2.6 Store
   sample at 4C
             8315A -  32
                             Revision  1
                          January  1995

-------
                                            METHOD  83ISA
                                              continued
                                                  74.1.1 Measure volume
                                                   of aqueous phase of
                                                 sample; pour sample into
                                                   separator/ funnel and
                                                 drain methytene chloride
                                                  (from Method 0011) into
                                                      volumetric flask
                                                  7.4.1.2 Extract aqueous
                                                  solution with methytene
                                                  chloride; add methytene
                                                    chloride extracts to
                                                      volumetric flask
                                                  7.4.1.3 Dilute to volume
                                                  with metiytene chloride;
                                                  mix well; remove aliquot
7.4.1.5 Store
sample at 4C
                                                                                  7 4.1.4 Dilute
                                                                                extract with mobile
                                                                                     phase
  7.4.1.4 Exchange
solvent with methanol
    as in 7.3.5.5
a high concentration
 of formaldehyde?
      7.4.1.4
    Does sample
     have a low
   concentration of
   formaldehyde?
7.4.1.4 Concentrate
  extract during
 solvent exchange
     process
                                             8315A  -  33
                                                                        Revision  1
                                                                     January  1995

-------
                                    METHOD  83ISA
                                      continued
7 5.2 Set LC conditions
to produce appropriate
    retention times
       75.1 Option
          or2LC
       conditions''
    7 5.1 Set LC
conditions to produce
appropriate retention
       times
   7.5.2.1 Filter and
  degas mobile phase
  7.6.2 Process calibration
  standards through same
processing steps as samples
                                    7.6.3 - 7.6.4
                                Analyze solvent blank
                              and calibration standards;
                                 tabulate peak areas
                               7.6.5 Determine response
                              factor at each concentration
                                      76.5
                                      Does
                                    calibration
                                   check meet
                                     criteria?
                                    O
                                       7.6.5 Prepare new
                                          calibration
                                          standards
                                    8315A  -  34
                                                        Revision  1
                                                     January  1995

-------
                                 METHOD 83ISA
                                   continued
                                 7.6.6 - 7.6.7 Verity
                             calibration curve every day;
                               reanalyze 1 calibration
                                 standard after 10
                                sample runs or less
                                 7.7 Analyze samples
                                     byHPLC
 7.7.2 Inject a smaller
volume or dilute sample
    7.7.4 Further
  cleanup is required
    7.7.2
  Does peak
 area exceed
  calibration
   curve'
  7.7.4 Are
interferences
  present?
7.8.1 Calculate each
response (actor, mean
response factor, and
percent RSD
i
1
7.8.2 - 7.8.5
Calculate analyte
concentrations
1
1
                                     Stop


                                  8315A -  35
                                               Revision 1
                                            January  1995

-------
                                  APPENDIX A

            RECRYSTALLIZATION OF 2,4-DINITROPHENYLHYDRAZINE (DNPH)


NOTE:  This  procedure  should  be  performed in  a  properly  ventilated  hood.
       Inhalation  of acetonitrile can  result  in  nose  and throat  irritation
       (brief  exposure  at  500  ppm)  or  more  serious  effects   at  higher
       concentration and/or longer exposures.

     A.I   Prepare  a  saturated solution of DNPH  by  boiling excess DNPH in 200 ml
of acetonitrile for approximately 1 hour.

     A.2   After 1  hour, remove and transfer the supernatant to a covered beaker
on  a hot  plate  and  allow gradual  cooling to  40  to  60"C.    Maintain  this
temperature range until 95% of the solvent has  evaporated,  leaving  crystals.

     A.3   Decant the solution  to  waste  and rinse  the remaining  crystals twice
with three times their apparent volume of acetonitrile.

     A.4   Transfer the  crystals to a clean beaker, add 200 ml  of acetonitrile,
heat to boiling,  and  again  let  the crystals  grow slowly at 40 to 60°C until 95%
of the solvent has evaporated.   Repeat the rinsing process as  in Sec. A.3.

     A.5   Take  an  aliquot  of  the   second  rinse, dilute   10  times  with
acetonitrile, acidify with  1  ml of 3.8  M perchloric acid per 100 ml  of DNPH
solution, and analyze with HPLC as in Sec. 7.0  for Procedure  2.   An acceptable
impurity level is less than 0.025 ng/juL of formaldehyde  in recrystallized DNPH
reagent or below the sensitivity (ppb, v/v) level  indicated in Table 3 for the
anticipated sample volume.

     A.6   If the impurity level is not satisfactory, pipet off the solution to
waste, repeat the  recrystallization  as  in  Sec. A.4  but  rinse with  two 25 mL
portions of acetonitrile.  Prep and analyze the second rinse  as  in  Sec. A.5.

     A.7   When the impurity level  is satisfactory, place  the crystals  in  an all-
glass reagent bottle, add another  25 ml of acetonitrile,  stopper, and shake the
bottle.  Use  clean pipets  when removing  the saturated DNPH stock  solution to
reduce the possibility of contamination of the solution. Maintain only a minimum
volume of the saturated solution adequate for day to day operation  to minimize
waste of the purified reagent.
                                  8315A - 36                        Revision 1
                                                                  January 1995

-------
                                 METHOD 8321A

                SOLVENT EXTRACTABLE NON-VOLATILE COMPOUNDS BY
     HIGH PERFORMANCE LIQUID CHROMATOGRAPHY/THERMOSPRAY/MASS SPECTROMETRY
                  (HPLC/TS/MS)  OR ULTRAVIOLET  (UV)  DETECTION
1.0   SCOPE AND APPLICATION

      1.1   This method covers the use of high performance liquid chromatography
(HPLC),  coupled with  either  thermospray-mass  spectrometry   (TS-MS),  and/or
ultraviolet (UV), for the determination of disperse azo dyes,  organophosphorus
compounds,  and  Tris-(2,3-dibromopropyl)phosphate,   chlorinated  phenoxyacid
compounds  and  their esters, and carbamates  in wastewater, ground  water,  and
soil/sediment matrices.  Data are also provided for chlorophenoxy acid herbicides
in fly  ash (Table  15), however, recoveries  for most compounds  are  very poor
indicating poor extraction  efficiency  for  these analytes  using the extraction
procedure  included   in  this method.    Additionally,   it  may  apply to  other
non-volatile compounds that are solvent extractable, are amenable to HPLC,  and
are ionizable under thermospray introduction for mass spectrometric detection.
The following compounds can be determined by this method:
      Compound Name
CAS No.'
Azo Dyes
Disperse Red 1
Disperse Red 5
Disperse Red 13
Disperse Yellow 5
Disperse Orange 3
Disperse Orange 30
Disperse Brown 1
Solvent Red 3
Solvent Red 23
Anthraquinone Dyes
Disperse Blue 3
Disperse Blue 14
Disperse Red 60
Coumarin Dyes
Fluorescent Briqhteners
Fluorescent Brightener 61
Fluorescent Brightener 236
Alkaloids
Caffeine
Strychnine

2872-52-8
3769-57-1
126038-78-6
6439-53-8
730-40-5
5261-31-4
17464-91-4
6535-42-8
85-86-9

2475-46-9
2475-44-7
17418-58-5


8066-05-5
3333-62-8

58-08-2
57-24-9
                                   8321A  -  1
         Revision 1
       January 1995

-------
Compound Name
CAS No."
Organophosphorus Compounds
Methomyl                                           16752-77-5
Thiofanox                                          39196-18-4
Famphur                                               52-85-7
Asulam                                             3337-71-1
Dichlorvos                                            62-73-7
Dimethoate                                            60-51-5
Disulfoton                                           298-04-4
Fensulfothion                                        115-90-2
Merphos                                              150-50-5
Methyl parathion                                     298-00-0
Monocrotophos                                        919-44-8
Naled                                                300-76-5
Phorate                                              298-02-2
Trichlorfon                                           52-68-6
Tris-(2,3-Dibromopropy1) phosphate, (Tris-BP)         126-72-7

Chlorinated Phenoxyacid Compounds
Dalapon                                               75-99-0
Dicamba                                            1918-00-9
2,4-D                                                 94-75-7
MCPA                                                  94-74-6
MCPP                                               7085-19-0
Dichlorprop                                          120-36-5
2,4,5-T                                               93-76-5
Silvex (2,4,5-TP)                                     93-72-1
Dinoseb                                               88-85-7
2,4-DB                                                94-82-6
2,4-D, butoxyethanol ester                         1929-73-3
2,4-D, ethylhexyl ester                            1928-43-4
2,4,5-T, butyl ester                                  93-79-8
2,4,5-T, butoxyethanol ester                       2545-59-7

Carbamates

Aldicarb*                                            116-06-3
Adicarb Sulfone                                    1646-88-4
Aldicarb Sulfoxide                                 1646-87-3
Aminocarb                                          2032-59-9
Barban                                               101-27-9
Benomyl                                            17804-35-2
Bromacil                                             314-40-9
Bendiocarb*                                        22781-23-3
Carbaryl*                                             63-25-2
Carbendazim*                                       10605-21-7
3-Hydroxy-Carbofuran                               16655-82-6
Carbofuran*                                        1563-66-2
                             8321A -  2
          Revision  1
        January  1995

-------
      Compound Name                                        CAS No."
      Carbamates (continued)

      Chloroxuron                                         1982-47-4
      Chloropropham                                        101-21-3
      Diuron                                               330-54-1
      Fenuron                                              101-42-8
      Fluometuron                                         2164-17-2
      Linuron*                                             330-55-2
      Methiocarb                                          2032-65-7
      Methomyl*                                          16752-77-5
      Mexacarbate                                          315-18-4
      Monuron                                              150-68-5
      Neburon                                              555-37-3
      Oxamyl*                                            23135-22-0
      Propachlor                                          1918-16-7
      Propham                                              122-42-9
      Propoxur                                             114-26-1
      Siduron                                             1982-49-6
      Tebuthiuron                                        34014-18-1
      "   Chemical Abstract Services Registry Number.
          These carbamates  were  tested in a multi-laboratory  evaluation;  all
          others were tested in a single-laboratory evaluation.


      1.2   This method may be applicable  to  the analysis of other non-volatile
or semi volatile compounds.

      1.3   Tris-BP  has  been classified  as  a  carcinogen.   Purified  standard
material and stock standard solutions should be handled in a hood.

      1.4   Method  8321   is  designed  to  detect  the chlorinated  phenoxyacid
compounds (free acid form)  and their  esters  without  the  use of hydrolysis  and
esterification in the extraction procedure.

      1.5   The compounds were chosen for  analysis  by HPLC/MS because they have
been designated as  problem compounds that are hard  to analyze by traditional
chromatographic methods  (e.g.  gas chromatography).   The sensitivity  of this
method is dependent  upon  the level of interferants within  a given matrix,  and
varies  with  compound   class  and even   with  compounds  within  that  class.
Additionally,   the limit  of detection  (LOD)  is  dependent  upon  the  mode  of
operation of the mass spectrometer.   For  example,  the  LOD  for  caffeine in  the
selected reaction monitoring  (SRM) mode  is  45  pg of standard  injected  (10  pi
injection),  while for Disperse Red 1  the  LOD is  180  pg.   The  LOD for  caffeine
under single quadrupole  scanning  is 84 pg and is 600 pg for Disperse Red 1 under
similar scanning conditions.


                                  8321A - 3                         Revision 1
                                                                  January 1995

-------
      1.6   The  experimentally  determined limits  of  detection  (LOD)  for the
target analytes are presented in Tables 3, 10, 13, and  14.  For further compound
identification, MS/MS (CAD -  Collision Activated Dissociation) can be used as an
optional extension of this method.

      1.7   This method  is restricted  to  use by,  or under the supervision of,
analysts experienced in the use of high performance liquid chromatographs/mass
spectrometers and skilled in  the interpretation of liquid chromatograms and mass
spectra.   Each  analyst  must demonstrate the  ability to generate  acceptable
results with this method.
2.0   SUMMARY OF METHOD

      2.1   This  method   provides   reverse   phase   high  performance  liquid
chromatographic (RP/HPLC) and thermospray (TS) mass spectrometric (MS) conditions
for the detection of the target analytes.  Quantitative analysis  is performed by
TS/MS, using an external  standard approach.  Sample extracts can be analyzed by
direct  injection  into the  thermospray  or  onto a   liquid  chromatographic-
thermospray interface.  A gradient elution program is  used on the chromatograph
to separate the  compounds.   Detection  is  achieved both by negative ionization
(discharge electrode)  and  positive ionization, with  a single  quadrupole mass
spectrometer.   Since  this method  is based  on  an HPLC technique, the  use of
ultraviolet (UV) detection is optional  on routine samples.

      2.2   Prior  to  the  use  of this  method,  appropriate  sample preparation
techniques must be used.

            2.2.1  Samples for  analysis of chlorinated phenoxyacid  compounds are
      prepared by a modification of Method 8151 (see Sec. 7.1.2).   In general,
      one  liter of  aqueous  sample  or  fifty grams  of  solid  sample  are  pH
      adjusted, extracted with  diethyl ether, concentrated and solvent exchanged
      to acetonitrile.

            2.2.2  Samples for  analysis of the  other target analytes are prepared
      by  established  extraction  techniques.    In  general,  water  samples  are
      extracted  at  a  neutral pH  with methylene chloride, using an appropriate
      3500 series  method.   An  appropriate 3500  series method  using  methylene
      chloride/acetone  (1:1)  is used  for solid samples.   A  micro-extraction
      technique  is  included for the extraction  of  Tris-BP from  aqueous  and
      non-aqueous matrices.

            2.2.3  For carbamates one liter aqueous samples or  forty grams of
      solid sample are methylene chloride extracted (refer to appropriate 3500
      series method),  concentrated  (preferably using  a rotary  evaporator with
      adapter) and solvent exchanged with methanol.

      2.3   An   optional   thermospray-mass   spectrometry/mass   spectrometry
(TS-MS/MS) confirmatory method is provided.   Confirmation is obtained by using
MS/MS Collision Activated  Dissociation  (CAD) or wire-repeller CAD.
                                   8321A -  4                        Revision 1
                                                                  January 1995

-------
3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, 8000 and 8151.

      3.2   The  use  of  Florisil  Column  Cleanup  (Method  3620)  has  been
demonstrated to yield recoveries  less than 85% for some  of the compounds in this
method, and is therefore not recommended for all compounds.  Refer to Table 2 of
Method  3620 for  recoveries  of  organophosphorus  compounds  as  a  function  of
Florisil fractions.

      3.3   Compounds with  high  proton affinity  may  mask some  of the target
analytes.  Therefore, an HPLC must be used as a chromatographic separator, for
quantitative analysis.

      3.4   Analytical difficulties encountered with specific organophosphorus
compounds, as applied in this method, may include (but are not limited to) the
following:

            3.4.1  Methyl  parathion shows  some minor degradation upon analysis.

            3.4.2  Naled can  undergo debromination to  form dichlorvos.

            3.4.3  Merphos often  contains contamination  from merphos  oxide.
      Oxidation  of  merphos   can  occur  during   storage,  and  possibly  upon
      introduction into the mass spectrometer.

            Refer to Method 8141 for other compound problems as related to the
      various extraction methods.

      3.5   The chlorinated phenoxy acid compounds, being  strong organic acids,
react  readily  with  alkaline  substances   and  may  be  lost  during  analysis.
Therefore, glassware and glass wool must be acid-rinsed, and sodium sulfate must
be acidified with sulfuric acid prior to use to avoid this possibility.

      3.6   Due to the reactivity of the chlorinated herbicides,  the standards
must be prepared in acetonitrile.  Methylation will occur slowly,  if prepared in
methanol.

      3.7  Benomyl  is known to quickly degrade to carbendazim in the environment
(Reference 21).

      3.8   Solvents, reagents, glassware, and other sample processing hardware
may  yield  discrete  artifacts  or  elevated  baselines,  or   both,   causing
misinterpretation of chromatograms or spectra.   All of these materials must be
demonstrated to be free  from  interferences under the conditions of  the analysis
by running reagent blanks.  Specific selection  of reagents and purification of
solvents by distillation in all-glass systems may be required.

      3.9   Interferants  co-extracted  from the sample  will  vary considerably
from source to source.  Retention times of target analytes must be verified by
using reference standards.
                                   8321A  -  5                         Revision 1
                                                                  January 1995

-------
      3.10  The optional use of HPLC/MS/MS methods aids in the confirmation of
specific analytes.  These methods  are  less  subject to chemical noise than other
mass spectrometric methods.


4.0   APPARATUS AND MATERIALS

      4.1   HPLC/MS

            4.1.1  High Performance  Liquid  Chromatograph (HPLC) - An analytical
      system  with  programmable  solvent   delivery  system  and  all  required
      accessories including 10-/zL injection  loop,  analytical  columns, purging
      gases, etc.  The  solvent delivery system must be capable, at a minimum, of
      a binary solvent system.  The chromatographic system  must  be capable of
      interfacing with a Mass Spectrometer (MS).

                   4.1.1.1     HPLC  Post-Column Addition Pump  - A pump for post
            column  addition  should  be used.   Ideally,  this pump  should  be a
            syringe  pump,  and  does  not  have  to  be  capable  of  solvent
            programming.

                   4.1.1.2     Recommended HPLC Columns  - A guard column and an
            analytical column are required.

                         4.1.1.2.1   Guard  Column  -  C18  reverse   phase  guard
                   column, 10 mm x 2.6 mm  ID,  0.5 jiim frit,  or equivalent.

                         4.1.1.2.2   Analytical   Column  -  C18  reverse  phase
                   column, 100 mm x 2 mm ID, 5  urn particle size of ODS-Hypersil;
                   or C8  reversed phase column,  100 mm x 2 mm ID,  3 /xm particle
                   size of MOS2-Hypersil,  or equivalent.

            4.1.2  HPLC/MS interface(s)

                   4.1.2.1     Micromixer  - 10-jxL, interfaces HPLC column system
            with HPLC  post-column addition solvent system.

                   4.1.2.2     Interface -  Thermospray ionization interface and
            source  that  will give  acceptable  calibration  response  for  each
            analyte of interest at the concentration required.  The source must
            be capable of generating both  positive and negative ions, and have
            a discharge  electrode or filament.

            4.1.3  Mass  spectrometer  system   -  A   single   quadrupole  mass
      spectrometer  capable of  scanning  from 1 to 1000  amu.   The  spectrometer
      must also be capable of scanning from 150 to 450 amu in 1.5 sec. or less,
      using 70  volts  (nominal)  electron   energy  in  the positive  or negative
      electron impact modes.  In addition,  the mass spectrometer must be capable
      of producing a calibrated  mass spectrum for PEG 400,  600, or 800  (see Sec.
      5.14).
                                   8321A -  6                         Revision 1
                                                                  January 1995

-------
                   4.1.3.1     Optional  triple quadrupole  mass  spectrometer -
            capable of generating daughter ion spectra with a collision gas in
            the second quadrupole and operation in the single quadrupole mode.

            4.1.4  Data System - A computer  system  that  allows  the continuous
      acquisition  and  storage on machine-readable  media of all  mass spectra
      obtained throughout the duration  of  the  chromatographic  program must be
      interfaced to the mass  spectrometer.  The computer must have software that
      allows any MS data file to be searched for ions of a specified mass, and
      such ion abundances to be plotted versus time or scan number.  This type
      of plot is defined as an Extracted Ion Current Profile (EICP).  Software
      must also be  available  that allows  integration  of  the  abundances in any
      EICP between specified  time or scan-number limits.  There must be computer
      software available  to operate the specific modes of  the mass spectrometer.

      4.2   HPLC  with  UV  detector  -  An  analytical  system  with  solvent
programmable  pumping  system  for  at  least  a binary  solvent  system,  and all
required  accessories   including  syringes,  10-juL  injection  loop,  analytical
columns, purging gases, etc.   An automatic injector is optional, but is useful
for multiple samples.   The columns specified  in Sec. 4.1.1.2 are also used with
this system.

            4.2.1  If   the  UV  detector  is  to  be  used  in  tandem with  the
      thermospray  interface,  then   the  detector   cell  must  be  capable  of
      withstanding high pressures (up to 6000  psi).   However,  the UV detector
      may be attached  to  an HPLC  independent  of the HPLC/TS/MS and in that case
      standard HPLC pressures are acceptable.

      4.3   Purification Equipment for Azo Dye Standards

            4.3.1  Soxhlet extraction  apparatus.

            4.3.2  Extraction thimbles,  single  thickness, 43  x  123  mm.

            4.3.3  Filter  paper,   9.0  cm   (Whatman   qualitative   No.   1  or
      equivalent).

            4.3.4  Silica-gel  column  -  3  in.  x 8 in., packed with  Silica gel
      (Type 60,  EM reagent 70/230 mesh).

      4.4   Extraction equipment for Chlorinated Phenoxyacid Compounds

            4.4.1  Erlenmeyer flasks  - 500-mL wide-mouth  Pyrex®,  500-mL Pyrex®,
      with 24/40 ground glass joint,  1000-mL Pyrex®.

            4.4.2  Separatory funnel  -  2000-mL.

            4.4.3  Graduated  cylinder  -  1000-mL.

            4.4.4  Funnel  - 75 mm diameter.

            4.4.5  Wrist  shaker  - Burrell  Model 75  or  equivalent.

            4.4.6  pH  meter.

                                  8321A -  7                         Revision 1
                                                                  January 1995

-------
      4.5   Kuderna-Danish (K-D) apparatus (optional).

            4.5.1  Concentrator tube -  10-mL graduated (Kontes K-570050-1025 or
      equivalent).  A ground  glass  stopper is used to  prevent  evaporation of
      extracts.

            4.5.2  Evaporation   flask -   500-mL  (Kontes   K-570001-500   or
      equivalent).   Attach  to  concentrator  tube  with  springs,  clamps,  or
      equivalent.

            4.5.3  Snyder column  -   Two-ball  micro  (Kontes K-569001-0219 or
      equivalent).

            4.5.4  Springs -   1/2 in.  (Kontes  K-662750 or  equivalent).

      NOTE: The  following glassware is recommended  for  the  purpose of solvent
            recovery during the  concentration procedures  requiring the use of
            Kuderna-Danish evaporative concentrators.  Incorporation  of this
            apparatus may be required by State  or local municipality regulations
            that govern air emissions  of volatile organics.  EPA recommends the
            incorporation of  this type of reclamation  system  as  a  method to
            implement an  emissions  reduction  program.  Solvent  recovery is a
            means to conform  with waste minimization  and  pollution prevention
            initiatives.

           4.5.5    Solvent vapor recovery  system  (Kontes  K-545000-1006  or K-
     547300-0000, Ace Glass 6614-30, or equivalent).

     4.6   Disposable serological  pipets -  5 ml x 1/10,  5.5 mm ID.

     4.7   Collection  tube -  15-mL conical,  graduated (Kimble  No. 45165 or
equivalent).

     4.8   Vials  - 5-mL  conical,  glass, with Teflon®-!ined screw-caps or crimp
tops.

     4.9   Glass  wool -  Supelco  No.  2-0411  or  equivalent.

     4.10  Microsyringes - 100-juL, 50-juL,  10-/A (Hamilton  701 N or equivalent),
and 50 nl  (Blunted, Hamilton  705SNR or equivalent).

     4.11  Rotary evaporator - Equipped with 1000-mL receiving flask.

     4.12  Balances  - Analytical,  0.0001 g, Top-loading, 0.01 g.

     4.13  Volumetric flasks,  Class  A  - 10-mL  to 1000-mL.

     4.14  Graduated cylinder  -  100-mL.

     4.15  Separatory funnel - 250-mL.

     4.16  Separatory funnel - 2-liter, with Teflon® stopcock.

     4.17  Concentrator  adaptor  (optional-  for carbamate extraction).

                                  8321A -  8                         Revision 1
                                                                  January  1995

-------
5.0   REAGENTS

      5.1   Reagent  grade chemicals shall be used in all  tests.   Unless otherwise
indicated, it is intended that all  reagents  shall  conform  to the specifications
of the Committee on Analytical Reagents of the American Chemical Society, where
such  specifications  are  available.   Other grades may  be  used, provided it  is
first ascertained that the reagent is of sufficiently high purity  to permit  its
use without lessening the accuracy of the determination.

      5.2   Organic  free  reagent water.   All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Sodium sulfate (granular, anhydrous),  Na2S04.  Purify  by heating  at
400°C for 4 hours  in a shallow tray, or by precleaning the sodium  sulfate with
methylene chloride.

      5.4   Ammonium acetate,  NH4OOCCH3,  solution (0.1  M).   Filter through a 0.45
micron membrane filter (Millipore HA or equivalent).

      5.5   Acetic acid,  CH3C02H

      5.6   Sulfuric acid  solution

           5.6.1    (1:1,  v/v) - Slowly add 50 ml H2S04 (sp. gr. 1.84) to 50  ml
      of water.

           5.6.2    (1:3,  v/v) - slowly add 25 ml H2S04 (sp. gr. 1.84) to 75  ml
      of water.

      5.7   Argon  gas,  99+% pure.

      5.8   Solvents

           5.8.1    Methylene  chloride, CH2C12 - Pesticide quality or equivalent.

           5.8.2    Toluene, C6H5CH3 - Pesticide quality or equivalent.

           5.8.3    Acetone, CH3COCH3 - Pesticide quality or equivalent.

           5.8.4    Diethyl  Ether,  C2H5OC2H5 -  Pesticide quality or equivalent.
      Must  be free of peroxides  as indicated  by test  strips (EM  Quant,   or
      equivalent).   Procedures for removal of  peroxides  are  provided  with the
      test  strips.   After  cleanup,  20 ml of  ethyl  alcohol  preservative must  be
      added to each liter  of  ether.

           5.8.5    Methanol,  CH3OH - HPLC  quality or  equivalent.

           5.8.6    Acetonitrile, CH3CN  - HPLC  quality or equivalent.

           5.8.7    Ethyl  acetate CH3C02C2H5 -  Pesticide  quality or  equivalent.
                                   8321A  -  9                         Revision 1
                                                                  January 1995

-------
      5.9   Standard  Materials  -  pure standard materials or certified solutions
of each analyte targeted for analysis.  Disperse azo dyes must be purified before
use according to Sec. 5.10.

      5.10  Disperse  Azo Dye Purification

           5.10.1  Two  procedures  are  involved.   The first step is the Soxhlet
      extraction  of  the dye for 24  hours with toluene  and evaporation  of the
      liquid  extract to dryness, using a  rotary  evaporator.   The solid is then
      recrystallized from toluene, and dried  in an oven at approximately 100°C.
      If  this step  does  not give the required  purity,  column  chromatography
      should  be  employed.   Load the  solid onto a 3  x  8 inch  silica gel  column
      (Sec.   4.3.4),  and   elute  with  diethyl  ether.    Separate  impurities
      chromatographically,  and collect the major dye fraction.

      5.11  Stock  standard  solutions  -   Can  be  prepared from  pure  standard
materials or  can  be purchased as certified  solutions.   Commercially-prepared
stock standards can  be  used if they  are verified against  EPA standards.  If EPA
standards are not available for verification, then  standards  certified  by the
manufacturer  and verified against  a standard  made  from pure  material  is
acceptable.

           5.11.1  Prepare  stock standard  solutions by accurately weighing 0.0100
      g  of  pure  material.   Dissolve  the material  in  methanol  or other suitable
      solvent  (e.g. prepare Tris-BP in  ethyl acetate), and dilute to known volume
      in a  volumetric flask.

NOTE:      Due to the reactivity of the  chlorinated  herbicides,  the standards
           must be prepared in acetonitrile.   Methylation will occur if prepared
           in  methanol.

           If compound purity  is certified at 96% or greater,  the weight can be
      used  without  correction to  calculate  the  concentration  of the  stock
      standard.    Commercially prepared  stock standards  can  be  used  at  any
      concentration   if  they  are  certified  by  the   manufacturer  or  by  an
      independent  source.

           5.11.2  Transfer the  stock  standard solutions into  glass vials with
      Teflon®-!ined  screw-caps or crimp-tops.   Store at 4"C  and  protect from
      light.   Stock  standard solutions should be  checked frequently for signs of
      degradation or evaporation, especially just prior to preparing calibration
      standards.

      5.12  Calibration  standards -  A  minimum of five  concentrations  for each
parameter of interest should be prepared through  dilution of the stock standards
with methanol (or other suitable solvent).  One of these concentrations should
be near, but above,  the MDL.  The  remaining concentrations should correspond to
the expected range of concentrations found in real  samples, or  should define the
working range of the HPLC-UV/VIS or HPLC-TS/MS.   Calibration  standards must be
replaced after one  or two months,  or sooner  if comparison with check standards
indicates a problem.

      5.13  Surrogate standards - The analyst should  monitor the performance of
the  extraction,  cleanup (when  used), and  analytical  system, along  with the

                                  8321A  - 10                        Revision 1
                                                                  January 1995

-------
effectiveness of the method in dealing with  each  sample matrix, by spiking each
sample, standard, and blank with one or two surrogates (e.g., organophosphorus
or chlorinated phenoxyacid compounds not expected to  be present in the sample).

      5.14  HPLC/MS tuning  standard  -  Polyethylene glycol 400  (PEG-400), PEG-600
or PEG-800.   Dilute to  10 percent  (v/v)  in methanol.   Dependent upon analyte
molecular weight range:  m.w. < 500 amu, use PEG-400; m.w. > 500 amu, use PEG-600,
or PEG-800.
6.0   SAMPLE  COLLECTION,  PRESERVATION, AND  HANDLING

      6.1   See  the  introductory  material  to  this Chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Sample  preparation - Samples for analysis of  disperse  azo dyes and
organophosphorus compounds must be prepared by an appropriate 3500 series method
prior to HPLC/MS analysis.:

      Samples  for  the analysis of Tris-(2,3-dibromopropyl)phosphate wastewater
must be prepared according to Sec. 7.1.1 prior to HPLC/MS  analysis.  Samples for
the analysis  of chlorinated phenoxyacid compounds and their esters  should be
prepared according to Sec. 7.1.2 prior to HPLC/MS analysis.

           7.1.1   Microextraction  for Tris-BP:

                   7.1.1.1  Solid Samples

                            7.1.1.1.1   Weigh  a 1-gram portion of  the sample
                   into  a tared beaker.   If the sample  appears moist,  add an
                   equivalent amount of anhydrous sodium  sulfate and mix well.
                   Add 100 jxL of Tris-BP (approximate concentration 1000 mg/L)
                   to the sample selected for  spiking;  the amount  added should
                   result in  a  final  concentration of 100 ng/juL  in  the  1-mL
                   extract.

                            7.1.1.1.2   Remove  the  glass  wool  plug  from  a
                   disposable serological  pipet.  Insert  a  1 cm plug of clean
                   silane treated glass wool to the bottom (narrow end) of the
                   pipet.   Pack 2 cm of anhydrous sodium sulfate  onto the top of
                   the glass wool.   Wash pipet  and contents   with 3  -  5 mL of
                   methanol.

                            7.1.1.1.3   Pack the sample into the pipet prepared
                   according to  Sec.  7.1.1.1.2.   If packing material has dried,
                   wet with a few mL of methanol first,  then pack  sample  into
                   the pipet.

                            7.1.1.1.4   Extract the sampl e wi th 3 mL of methanol
                   followed by 4 mL of 50% (v/v)  methanol/methylene  chloride
                   (rinse the  sample beaker  with  each  volume of  extraction

                                  8321A - 11                        Revision 1
                                                                  January 1995

-------
                   solvent  prior  to  adding  it  to  the  pipet containing  the
                   sample).   Collect  the extract in  a 15-mL  graduated glass
                   tube.

                            7.1.1.1.5    Evaporate the extract to  1 mL using the
                   nitrogen  blowdown  technique  (Sec.  7.1.1.1.6).   Record  the
                   volume.   It  may not  be  possible to evaporate  some sludge
                   samples to  a  reasonable  concentration.

                            7.1.1.1.6    Nitrogen Blowdown Technique

                                         7.1.1.1.6.1Place the concentrator tube
                            in  a   warm  water  bath   (approximately  35°C)  and
                            evaporate the solvent volume to the required level
                            using  a  gentle  stream of  clean,  dry  nitrogen
                            (filtered through a column  of activated carbon).

CAUTION:   Do not use plasticized tubing between the  carbon trap and the sample.

                                         7.1.1.1.6.2The  internal   wall   of  the
                            tube  must  be  rinsed  down  several  times  with
                            methylene chloride  during  the operation.   During
                            evaporation, the solvent level  in the tube must be
                            positioned to prevent water from condensing  into the
                            sample (i.e., the solvent level should be below the
                            level  of the water bath).Under   normal   operating
                            conditions,  the  extract should not  be  allowed to
                            become dry.  Proceed to  Sec.  7.1.1.1.7.

                            7.1.1.1.7    Transfer  the  extract to  a  glass vial
                   with  a  Teflon®-!ined  screw-cap or  crimp-top  and  store
                   refrigerated  at 4°C.   Proceed with HPLC  analysis.

                            7.1.1.1.8    Determination  of percent dry weight -
                   In  certain  cases,  sample results  are desired based on a dry
                   weight basis.   When  such data are  desired, or required,  a
                   portion  of  sample  for this  determination  should  be weighed
                   out at the  same  time as the  portion  used for  analytical
                   determination.

WARNING:   The drying  oven should be contained in a hood or vented.  Significant
           laboratory contamination may result from drying a heavily contaminated
           hazardous waste sample.

                            7.1.1.1.9    Immediately after  weighing  the sample
                   for extraction,  weigh 5-10  g  of the  sample  into  a tared
                   crucible.  Determine  the % dry weight of the  sample by drying
                   overnight  at 105°C.   Allow  to cool  in  a  desiccator before
                   weighing:


                            % dry weight = q of dry  sample x 100
                                             g  of sample


                                   8321A  - 12                        Revision 1
                                                                  January 1995

-------
                   7.1.1.2  Aqueous Samples

                            7.1.1.2.1   Using  a  100-mL  graduated  cylinder,
                   measure  100  ml  of  sample  and  transfer  it  to  a  250-mL
                   separatory  funnel.   Add  200  /xL  of Tris-BP  (approximate
                   concentration  1000 mg/L) to the sample selected for spiking;
                   the  amount  added  should result in  a  final  concentration of
                   200  ng//iL in  the 1-mL extract.

                            7.1.1.2.2   Add 10 ml of methylene chloride to the
                   separatory  funnel.   Seal  and  shake  the  separatory  funnel
                   three   times,   approximately  30  seconds  each  time,  with
                   periodic  venting to release excess  pressure.

NOTE:     Methylene  chloride  creates  excessive  pressure  rapidly;  therefore,
          initial venting should be done immediately after the separatory funnel
          has  been  sealed and shaken once.   Methylene chloride is a suspected
          carcinogen,  use necessary safety precautions.

                            7.1.1.2.3   Allow the organic layer to  separate from
                   the  water phase  for a minimum  of  10  minutes.   If the emulsion
                   interface between layers  is more  than one-third the size of
                   the  solvent   layer,  the  analyst  must  employ  mechanical
                   techniques  to  complete phase  separation.   See  Section 7.5,
                   Method  3510.

                            7.1.1.2.4   Collect the  extract in a 15-mL graduated
                   glass tube.  Proceed  as  in Sec. 7.1.1.1.5.

          7.1.2    Extraction for chlorinated phenoxyacid compounds  - Preparation
     of soil,  sediment, and other  solid samples  must follow Method 8151, with
     the exception of no hydrolysis or esterification.   (However, if the analyst
     desires  to  determine all   of the  phenoxyacid  moieties as  the  acid,
     hydrolysis may be performed.)   Section 7.1.2.1 presents an outline of the
     procedure  with the  appropriate changes necessary for determination  by
     Method  8321.    Section 7.1.2.2 describes  the  extraction procedure  for
     aqueous samples.

                   7.1.2.1  Extraction of  solid  samples

                            7.1.2.1.1   Add 50 g of soil/sediment sample to a
                   500-mL, wide   mouth  Erlenmeyer.   Add spiking  solutions  if
                   required, mix well and allow to stand  for  15 minutes.  Add 50
                   ml of   organic-free reagent water and stir  for  30  minutes.
                   Determine the  pH of the sample with a glass electrode and pH
                   meter,  while  stirring.   Adjust the pH to 2 with cold H2S04
                   (1:1) and monitor the pH  for  15 minutes,  with stirring.   If
                   necessary, add additional H2S04 until  the pH remains at 2.

                            7.1.2.1.2   Add 20 ml of acetone to the flask,  and
                   mix  the contents with the wrist shaker for 20 minutes.   Add
                   80 ml of diethyl ether to  the same flask, and shake again for
                                  8321A - 13                        Revision 1
                                                                  January 1995

-------
                   20  minutes.   Decant  the  extract and measure the  volume of
                   solvent  recovered.

                            7.1.2.1.3   Extract the sample twice more using 20
                   ml  of acetone  followed  by 80 ml  of diethyl  ether.   After
                   addition of  each  solvent,  the mixture should  be shaken with
                   the wrist shaker for 10 minutes and the acetone-ether extract
                   decanted.

                            7.1.2.1.4   After the third extraction, the volume
                   of extract recovered  should be at least 75% of the volume of
                   added  solvent.     If this   is  not  the  case,  additional
                   extractions  may be necessary.   Combine the extracts in a 2000
                   ml separatory funnel containing 250 ml of 5% acidified sodium
                   sulfate.   If an emulsion  forms,  slowly add  5  g  of acidified
                   sodium sulfate  (anhydrous) until  the solvent-water  mixture
                   separates.   A quantity of acidified  sodium  sulfate equal to
                   the weight of the  sample  may  be  added,  if necessary.

                            7.1.2.1.5   Check the pH of the extract.  If it is
                   not  at or below pH  2,  add more concentrated HC1  until  the
                   extract  is  stabilized at  the desired  pH.   Gently  mix  the
                   contents of  the separatory funnel for  1 minute and allow the
                   layers to separate.  Collect the  aqueous phase  in  a clean
                   beaker,  and  the   extract  phase  (top  layer)  in  a   500  ml
                   ground-glass Erlenmeyer flask.  Place the aqueous phase back
                   into  the separatory  funnel  and re-extract  using 25  ml of
                   diethyl  ether.  Allow the  layers  to separate and discard the
                   aqueous  layer.  Combine  the  ether  extracts  in  the  500 ml
                   Erlenmeyer flask.

                            7.1.2.1.6   Add 45 - 50 g  acidified anhydrous sodium
                   sulfate  to the combined ether  extracts. Allow the extract to
                   remain in contact  with the sodium  sulfate for approximately
                   2 hours.

NOTE:     The drying step is very  critical.  Any moisture remaining  in the ether
          will  result in low recoveries.  The amount of sodium sulfate used is
          adequate if some  free flowing crystals are visible when swirling the
          flask.   If all of the sodium  sulfate  solidifies in a cake, add a few
          additional  grams  of acidified  sodium  sulfate and  again  test  by
          swirling. The  2  hour drying time is a minimum; however,  the extracts
          may be  held overnight in contact with the  sodium sulfate.

                            7.1.2.1.7   Transfer the ether  extract,  through a
                   funnel plugged with acid-washed glass wool,  into  a 500-mL K-D
                   flask  equipped  with a 10  ml concentrator tube.   Use a glass
                   rod to crush caked sodium sulfate during the transfer.  Rinse
                   the  Erlenmeyer flask and column  with 20-30  ml  of  diethyl
                   ether  to complete  the  quantitative  transfer.    Reduce  the
                   volume of the  extract using  the macro K-D technique (Sec.
                   7.1.2.1.8).
                                  8321A - 14                        Revision 1
                                                                  January 1995

-------
          7.1.2.1.8    Add one or two clean boiling  chips  to
the  flask  and  attach  a three  ball  macro-Snyder  column.
Attach the  solvent  vapor recovery glassware  (condenser and
collection device)  (Sec. 4.5.5) to the Snyder column  of the
K-D apparatus following manufacturer's  instructions.  Prewet
the Snyder  column by  adding  about 1  ml of diethyl  ether  to
the top.   Place the  apparatus on  a hot  water  bath  (600-65°C)
so that the  concentrator tube is partially  immersed  in the
hot water and the entire lower rounded surface of  the flask
is bathed in vapor.   Adjust the  vertical  position  of the
apparatus and the  water temperature, as required,  to complete
the concentration in  15-20 minutes.   At the proper rate  of
distillation the balls of the column will  actively  chatter,
but the chambers will  not flood.  When the apparent volume  of
liquid reaches 5 ml, remove the  K-D apparatus from the water
bath and allow it  to drain and cool for at  least  10  minutes.

          7.1.2.1.9    Exchange the solvent  of the  extract  to
acetonitrile by quantitatively transferring the extract with
acetonitrile to a blow-down apparatus.  Add  a total of 5  ml
acetonitrile.   Reduce  the  extract volume according to Sec.
7.1.1.1.6, and adjust the final  volume to 1 ml.

7.1.2.2   Preparation  of  aqueous samples

          7.1.2.2.1    Using   a  1000-mL  graduated  cylinder,
measure 1  liter (nominal) of sample, record the sample  volume
to the nearest 5 ml, and transfer it  to a separatory funnel.
If high concentrations are anticipated,  a smaller  volume may
be used and then diluted with organic-free reagent  water  to
1 liter.  Adjust the pH  to  less than  2  with sulfuric acid
(1:1).

          7.1.2.2.2    Add 150 ml of  diethyl ether to the
sample bottle,  seal,  and shake  for 30  seconds to rinse the
walls. Transfer the  solvent wash to the separatory funnel and
extract the sample by shaking the funnel for 2 minutes with
periodic  venting  to  release excess  pressure.    Allow the
organic layer to separate from the water  layer for a minimum
of 10 minutes.  If the emulsion  interface between layers  is
more  than one-third  the size  of the solvent  layer, the
analyst must  employ mechanical  techniques  to complete the
phase  separation.  The  optimum  technique  depends  upon the
sample, and may include stirring,  filtration  of the  emulsion
through  glass  wool,  centrifugation,  or   other   physical
methods.   Drain the aqueous phase into a 1000-mL  Erlenmeyer
flask.

          7.1.2.2.3    Repeat  the extraction  two  more  times
using  100 ml  of  diethyl  ether  each  time.   Combine the
extracts  in a  500 ml  Erlenmeyer  flask.   (Rinse the 1000-mL
flask with each additional aliquot of extracting  solvent  to
make a quantitative  transfer.)
               8321A  - 15                        Revision 1
                                               January  1995

-------
                            7.1.2.2.4   Proceed to Sec. 7.1.2.1.6 (drying, K-D
                   concentration,    solvent   exchange,   and   final   volume
                   adjustment).

           7.1.3    Extraction for carbamates -  Preparation of soil, sediment, and
     other solid  samples  must follow an appropriate 3500  series method.

                   7.1.3.1  Forty gram quantities  are extracted with methylene
           chloride using  an appropriate 3500  series method.

                   7.1.3.2  Concentration  steps can be achieved using a rotary
           evaporator  or K-D, to  5-10 ml volumes.

                   7.1.3.3  Final  concentration  and  solvent  exchange  to 1-mL
           final volume of methanol,  can be done preferably using an adaptor on
           the  rotary evaporator.    If  an adaptor  is unavailable,  the  final
           concentration can be  achieved  using a gentle stream of nitrogen, in
           a fume  hood.

           7.1.4    Extraction  for carbamates  -  Preparation  of aqueous samples
     must  follow  an appropriate  3500 series method.

                   7.1.4.1  One liter quantities  are  extracted with methylene
           chloride using  an appropriate 3500  series method.

                   7.1.4.2  Final  concentration  and exchange to methanol is the
           same as applied in Sees.  7.1.3.2  and  7.1.3.3.

     7.2   Prior to HPLC analysis, the extraction  solvent must be exchanged to
methanol or acetonitrile (Sec.  7.1.2.1.9).  The  exchange  is performed using the
K-D procedures listed in all of the extraction methods.

     7.3   HPLC Chromatographic Conditions:

           7.3.1    Analyte-specific chromatographic conditions are shown in Table
     1.    Chromatographic conditions which  are  not  analyte-specific are as
     follows:

           Flow rate:                    0.4 mL/min
           Post-column mobile phase:     0.1 M ammonium acetate (1% methanol)
                                        (0.1 M ammonium acetate for phenoxyacid
                                        compounds)
           Post-column flow  rate:        0.8 mL/min

           7.3.2    If  there is a chromatographic problem from compound retention
     when  analyzing  for  disperse azo  dyes,  organophosphorus  compounds,  and
     Tris-(2,3-dibromopropyl)phosphate, a 2% constant flow of methylene chloride
     may be applied as  needed.   Methylene chloride/aqueous methanol solutions
     must  be used  with caution  as  HPLC  eluants.   Acetic  acid  (1%),  another
     mobile phase modifier, can  be  used  with compounds  with acid functional
     groups.

           7.3.3    A  total  flow   rate  of  1.0  to  1.5  mL/min  is necessary to
     maintain thermospray ionization.

                                  8321A - 16                        Revision 1
                                                                  January 1995

-------
     7.3.4   Retention  times   for   organophosphorus   compounds   on  the
specified analytical column are presented in Table 9.

7.4  Recommended HPLC/Thermospray/MS operating conditions:

     7.4.1   Positive lonization mode

     Repeller (wire or plate, optional):170 to 250 v  (sensitivity
           optimized). See Figure 2  for schematic of  source with  wire
           repeller.

     Discharge electrode:          off
     Filament:                     on or off (optional, analyte dependent)
     Mass range:                   150 to  450  amu  (analyte  dependent,
                                   expect  1  to   18  amu  higher   than
                                   molecular weight  of the compound).
     Scan time:                    1.50 sec/scan.

     7.4.2   Negative lonization mode

     Discharge electrode:          on
     Filament:                     off
     Mass Range:                   135 to 450 amu
     Scan time:                    1.50 sec/scan.

     7.4.3   Thermospray temperatures:

     Vaporizer control:            110°C  to 130°C.
     Vaporizer tip:                200eC  to 215eC.
     Jet:                          210°C  to 220°C.
     Source block:                 230°C  to  265°C.   (Some  compounds  may
                                   degrade  in the source block at higher
                                   temperatures,   operator   should   use
                                   knowledge  of  chemical  properties  to
                                   estimate proper source temperature).

     7.4.4   Sample  injection  volume: 20  juL  is  necessary  in order  to
overfill the 10-jiL injection loop.  If solids are  present in  the  extract,
allow them to settle or centrifuge the extract and withdraw the injection
volume from the clear layer.

7.5  Calibration:

     7.5.1   Thermospray/MS system - Must be hardware-tuned on quadrupole
1 (and quadrupole 3 for triple  quadrupoles)  for  accurate  mass  assignment,
sensitivity,  and  resolution.   This is  accomplished using  polyethylene
glycol (PEG) 400, 600, or 800 (see Sec. 5.14)  which  has average molecular
weights of 400, 600, and 800, respectively.  A mixture of  these  PEGs can
be made such that it will  approximate the expected working  mass range for
the  analyses.   Use  PEG  400  for  analysis  of chlorinated   phenoxyacid
compounds.    The  PEG  is  introduced via  the  thermospray   interface,
circumventing the HPLC.
                             8321A -  17                         Revision  1
                                                             January  1995

-------
        7.5.1.1  The mass calibration parameters are as follows:

        for PEG 400 and 600           for  PEG  800

        Mass range: 15 to 765 amu     Mass range:  15 to 900 amu
        Scan time: 5.00 sec/scan      Scan time: 5.00 sec/scan

        Approximately  100  scans  should  be acquired,  with 2  to 3
injections made.   The  scan  with the  best  fit to  the accurate  mass
table (see Tables 7 and 8)  should be used  as the calibration table.

        7.5.1.2  The low mass range from 15 to 100  amu  is  covered by
the ions  from  the ammonium acetate buffer  used in the thermospray
process:  NH4+  (18 amu),  NH4+H20 (36), CH3OHNH + (50) (methanol),  or
CH3CNNH4+  (59)  (acetonitrile),  and  CH3COOHNH4  (78) (acetic  acid).
The appearance  of the  m/z  50  or  59 ion  depends  upon the use  of
methanol or acetonitrile as  the organic modifier.   The higher  mass
range is covered by the ammonium ion adducts of the  various  ethylene
glycols (e.g. H(OCH2CH2)nOH where n=4, gives the H(OCH2CH2)4OHNH/ ion
at m/z 212).

7.5.2   Liquid Chromatograph

        7.5.2.1   Prepare calibration standards as outlined in Section
5.12.

        7.5.2.2   Choose the proper ionization conditions, as outlined
in Sec. 7.4.  Inject each calibration standard  onto the HPLC,  using
the chromatographic conditions  outlined in  Table  1.  Calculate  the
area under the curve for the mass chromatogram  of each  quantitation
ion.  For  example, Table 9  lists the retention times and the  major
ions  (>5%)  present in  the  positive  ionization thermospray  single
quadrupole spectra of the organophosphorus compounds.   In  most cases
the (M+H)+ and (M+NH4) + adduct  ions are the only ions of significant
abundance.   Plot  these ions  as area  response versus  the  amount
injected.   The  points  should  fall  on  a straight  line,  with a
correlation  coefficient of  at  least 0.99  (0.97  for chlorinated
phenoxyacid analytes).

        7.5.2.3   If HPLC-UV  detection is  also being used,  calibrate
the instrument  by preparing calibration standards as outlined in Sec.
5.12,  and injecting each calibration standard onto the HPLC using the
chromatographic conditions outlined in Table  1.  Integrate  the area
under   the   full   chromatographic  peak  for   each  concentration.
Quantitation by HPLC-UV may be  preferred  if it  is known that  sample
interference and/or analyte coelution are not a problem.

        7.5.2.4   For  the methods  specified  in Sees.  7.5.2.2  and
7.5.2.3,  the  retention time  of  the chromatographic  peak  is  an
important variable in analyte  identification.   Therefore,  the ratio
of the  retention time of the sample analyte to  the  standard analyte
should be 1.0 - 0.1.
                        8321A -  18                        Revision  1
                                                        January 1995

-------
             7.5.2.5   The  concentration of  the sample  analyte will  be
     determined by using the calibration curves determined in Sees. 7.5.2.2
     and 7.5.2.3.  These calibration curves must be generated on the same
     day as  each  sample is analyzed.  At  least  duplicate  determinations
     should be made for each sample extract.  Samples whose  concentrations
     exceed the standard calibration range should be diluted to fall within
     the range.

             7.5.2.6   Refer  to Method  8000  for  further  information  on
     calculations.

             7.5.2.7   Precision can  also be  calculated  from  the ratio  of
     response  (area)   to  the   amount  injected;   this  is defined as  the
     calibration  factor (CF)   for  each  standard concentration.    If  the
     percent relative standard deviation (%RSD)  of the CF is  less than  20
     percent over the  working  range, linearity  through  the origin  can  be
     assumed, and the  average  calibration  factor can  be  used  in place  of
     a calibration curve.  The CF and %RSD can be calculated as follows:

             CF = Total Area of Peak/Mass  injected (ng)

             %RSD = SD/CF x 100

     where:

             SD = Standard deviation between CFs

             CF = Average CF

7.6  Sample Analysis

     7.6.1   Once the LC/MS system has been calibrated as outlined in Sec.
7.5, then  it  is  ready for sample  analysis.   It is recommended  that  the
samples be  initially  analyzed  in the negative  ionization  mode.   If  low
levels of compounds are suspected then the samples should also be screened
in the positive ionization mode.

             7.6.1.1   A blank  20-jiL injection (methanol) must be analyzed
     after the standard(s)  analyses,  in order to determine any residual
     contamination of the Thermospray/HPLC/MS system.

             7.6.1.2   Take a 20 juL aliquot  of the sample extract from Sec.
     7.4.4.  Start the HPLC gradient elution, load and  inject  the  sample
     aliquot, and start the mass spectrometer data system analysis.

7.7  Calculations

     7.7.1   Using  the external  standard  calibration procedure  (Method
8000),  determine the identity  and quantity of each component peak  in the
sample reconstructed ion chromatogram which  corresponds to the  compounds
used for calibration processes.  See Method 8000 for calculation  equations.

     7.7.2   The retention time of the chromatographic peak is an important
parameter  for  the  identity  of  the  analyte.    However,  because  matrix

                            8321A  -  19                         Revision  1
                                                            January 1995

-------
interferences can change chromatographic column  conditions,  the  retention
times  are  not  as  significant,  and  the mass spectra  confirmations  are
important criteria for analyte identification.

7.8  Optional Thermospray HPLC/MS/MS confirmation

     7.8.1   With respect to  this  method,  MS/MS shall  be defined  as  the
daughter ion collision activated dissociation  acquisition  with quadrupole
one set on one mass (parent ion), quadrupole two pressurized with argon and
with a higher offset  voltage than normal, and quadrupole three set to scan
desired mass range.

     7.8.2   Since the thermospray process often generates only one or two
ions per compound, the use of MS/MS  is  a more specific mode  of  operation
yielding molecular structural information.  In this mode, fast  screening
of samples can be accomplished through direct injection of the sample into
the thermospray.

     7.8.3   For MS/MS experiments, the first  quadrupole  should  be set to
the protonated molecule or ammoniated adduct of the analyte  of  interest.
The third quadrupole should be  set to scan from 30 amu to just  above the
mass region of the protonated molecule.

     7.8.4   The  collision gas  pressure (Ar)  should  be set  at  about  1.0
mTorr and the collision energy at 20  eV.  If these parameters fail to give
considerable  fragmentation,  they  may be raised  above  these settings  to
create more and stronger collisions.

     7.8.5   For analytical  determinations, the base peak of the  collision
spectrum shall be taken as the quantification ion.  For  extra specificity,
a second ion should be chosen as a backup quantification ion.

     7.8.6   Generate  a calibration  curve as outlined in Sec. 7.5.2.

     7.8.7   For  analytical  determinations,  calibration  blanks must  be
analyzed in the MS/MS mode to determine  specific ion interferences.  If no
calibration  blanks  are  available,  chromatographic  separation  must  be
performed  to  assure   no  interferences  at  specific masses.   For  fast
screening,  the  MS/MS  spectra of  the standard and the  analyte  could  be
compared and the  ratios of the  three major (most  intense)  ions  examined.
These  ratios  should  be  approximately the  same  unless  there  is  an
interference.  If an interference appears, chromatography must be utilized.

     7.8.8   For unknown concentrations, the total  area of the quantitation
ion(s)  is calculated and the calibration curves generated as in Sec.  7.5
are used to attain an  injected weight number.

     7.8.9   MS/MS  techniques  can  also be  used  to  perform  structural
analysis on ions represented by unassigned m/z ratios.   The  procedure for
compounds of unknown structures is to  set up  a CAD experiment on the ion
of interest.  The spectrum generated  from this experiment will reflect the
structure of  the  compound  by its fragmentation  pattern.  A  trained mass
spectroscopist  and  some  history  of the sample  are  usually  needed  to
interpret  the  spectrum.  (CAD  experiments  on  actual  standards of  the

                             8321A -  20                        Revision 1
                                                             January 1995

-------
     expected  compound  are  necessary  for  confirmation  or  denial  of  that
     substance.)

     7.9  Optional wire-repeller  CAD confirmation

          7.9.1    See  Figure 3 for the  correct  position of the wire-repeller in
     the thermospray source block.

          7.9.2    Once the wire-repeller is inserted into the thermospray flow,
     the voltage can be increased to approximately  500 - 700 v.  Enough voltage
     is necessary to create fragment ions, but not so much that shorting occurs.

          7.9.3    Continue  as  outlined  in  Section  7.6.
8.0  QUALITY CONTROL

     8.1  Refer to Chapter One and Method 8000 for specific quality control (QC)
procedures.  Each  laboratory should maintain a formal quality assurance program.
The laboratory should also maintain  records  to document the quality of the data
generated.

     8.2   Quality  control  procedures necessary  to  evaluate the  HPLC  system
operation are found  in Method 8000, Sec. 7.0 and includes evaluation of retention
time windows, calibration verification and chromatographic analysis of samples.
Check the performance of the  entire  analytical system daily using data gathered
from analyses of blanks, standards,  and replicate samples.

           8.2.1    See Sec. 7.5.2.7 for required  HPLC/MS parameters for standard
     calibration  curve  %RSD  limits.

           8.2.2    See Sec. 7.5.2.4  regarding  retention time window QC limits.

           8.2.3    If any of the  chromatographic QC  limits  are not met,  the
     analyst should  examine the LC  system for:

           o        Leaks,
           o        Proper pressure delivery,
           o        A dirty guard column; may need  replacing or repacking, and
           o        Possible partial  thermospray  plugging.

           Any  of  the above  items will  necessitate shutting down  the  HPLC/TS
     system,  making  repairs and/or  replacements,  and  then  restarting  the
     analyses.  The calibration  standard should  be reanalyzed before any sample
     analyses, as described  in Sec. 7.5.

           8.2.4     The    experience   of   the   analyst   performing   liquid
     chromatography  is  invaluable to the success of the method.   Each day that
     analysis  is performed, the daily calibration standard should be  evaluated
     to determine if the chromatographic system is operating properly.   If any
     changes are  made to the system  (e.g.  column change), the  system must  be
     recalibrated.
                                  8321A - 21                        Revision 1
                                                                  January 1995

-------
     8.3     Initial  Demonstration  of  Proficiency  -  Each  laboratory  must
demonstrate initial proficiency with each sample preparation and determinative
method combination it utilizes, by  generating  data  of acceptable accuracy and
precision for target analytes in a  clean matrix.  The  laboratory must also repeat
the following operations whenever new staff are trained or significant changes
in instrumentation are made.  See Method 8000,  Sec.  8.0 for information on how
to accomplish this demonstration.

     8.4   Sample Quality Control for Preparation and Analysis - The laboratory
must also  have  procedures  for  documenting the effect of  the matrix  on method
performance  (precision,  accuracy,  and detection limit).   At a  minimum,  this
includes the analysis of QC samples  including  a  method  blank,  matrix spike,  a
duplicate, and a laboratory control sample (LCS) in each analytical  batch and the
addition of surrogates to each field sample and QC  sample.

           8.4.1    Documenting  the  effect of  the  matrix  should  include  the
     analysis of at least one matrix spike and  one duplicate unspiked sample or
     one matrix spike/matrix spike duplicate pair,   the decision on whether to
     prepare  and  analyze  duplicate  samples  or a  matrix  spike/matrix  spike
     duplicate must be based on a knowledge of  the samples in the sample batch.
     If samples are expected to contain target analytes, then laboratories may
     use one matrix spike and a duplicate analysis of an unspiked field sample.
     If samples are not expected to contain target analytes, laboratories should
     use a matrix  spike  and matrix  spike  duplicate pair.

           8.4.2  A Laboratory Control Sample (LCS) should  be included with each
     analytical  batch.   The LCS  consists of an aliquot  of a  clean  (control)
     matrix  similar to the sample matrix and of the  same weight or volume.  The
     LCS  is  spiked with the same  analytes at the same concentrations  as the
     matrix  spike.   When the results of  the matrix  spike analysis indicate  a
     potential problem due to the  sample matrix itself, the LCS results are used
     to verify  that the laboratory can perform the analysis in a clean matrix.

           8.4.3   See Method 8000,  Sec.   8.0 for the details on  carrying out
     sample  quality control procedures for  preparation and  analysis.

     8.5  Surrogate recoveries  - The laboratory must evaluate surrogate recovery
data from individual samples versus the surrogate control limits developed by the
laboratory.  See Method 8000, Sec. 8.0 for information on evaluating surrogate
data and developing and updating surrogate limits.

     8.6   It  is  recommended  that  the  laboratory   adopt  additional  quality
assurance practices for use with this method.  The  specific practices that are
most productive depend upon the needs of the  laboratory  and the nature of the
samples.   Whenever possible, the  laboratory should  analyze standard reference
materials and participate in relevant performance evaluation studies.


9.0  METHOD  PERFORMANCE

     9.1   Single  operator  accuracy and precision studies  have  been conducted
using spiked sediment, wastewater, sludge, and water samples.   The results are
presented in Tables 4, 5,  6,  11,  12,  15,  20 and 21.  Tables 4, 5, and 6 provide
single-laboratory  data  for  Disperse Red 1,   Table  11 with  organophosphorus

                                  8321A - 22                        Revision  1
                                                                  January 1995

-------
pesticides, Table 12 with Tris-BP,  Table 15 with chlorophenoxyacid herbicides and
Tables 20 and 21 with carbamates.

      9.2   LODs  should be calculated for the known analytes, on each instrument
to be used.  Tables 3, 10,  and 13 list limits of detection  (LOD) and/or estimated
quantitation limits (EQL) that are typical with this method.

           9.2.1    The LODs presented in this method were calculated by analyzing
      three  replicates  of   four  standard  concentrations,  with  the  lowest
      concentration  being  near the  instrument  detection  limit.    A  linear
      regression  was performed  on the  data  set  to  calculate  the  slope  and
      intercept.  Three times the standard  deviation  (3a)  of the lowest standard
      amount, along with the calculated slope and intercept, was used to find the
      LOD.   The  LOD was not calculated using the specifications in Chapter One,
      but  according to the ACS guidelines  specified  in  Reference 4.

           9.2.2    Table  17  presents  a  comparison  of the  LODs from Method 8150
      and  Method  8321 for the chlorinated  phenoxyacid  compounds.

      9.3   Table  16 presents multi-laboratory accuracy and precision data for the
chlorinated phenoxyacid herbicides.  The data summary is based on data from three
laboratories that  analyzed  duplicate solvent  solutions  at  each  concentration
specified in the Table.

      9.4   Tables 22  and 23 present the multi-laboratory  accuracy and precision
data  for   the  carbamates.    The  data  summary is  based  on data  from  nine
laboratories that analyzed triplicate solvent solutions  at each  concentration
level specified  in the Tables.


10.0  REFERENCES

1.    Voyksner,  R.D., Haney,  C.A., "Optimization and Application of Thermospray
      High-Performance  Liquid Chromatography/Mass  Spectrometry", Anal.  Chem..
      1985,  57,  991-996.

2.    Blakley,   C.R.,  Vestal,   M.L.,   "Thermospray   Interface   for   Liquid
      Chromatography/Mass Spectrometry", Anal.  Chem..  1983,  5_5, 750-754.

3.    Taylor, V.,  Hickey, D.M., Marsden, P.O.,  "Single Laboratory Validation of
      EPA  Method  8140", EPA-600/4-87/009, U.S.  Environmental Protection Agency,
      Las  Vegas,  NV,  1987, 144 pp.

4.    "Guidelines  for   Data  Acquisition  and  Data   Quality  Evaluation   in
      Environmental Chemistry", Anal. Chem., 1980, 52,  2242-2249.

5.    Betowski,  L.D., Jones,  T.L.,  "The  Analysis of Organophosphorus  Pesticide
      Samples by  HPLC/MS and  HPLC/MS/MS", Environmental Science and Technology,
      1988.

8.    U.S.  EPA:   2nd Annual  Report on Carcinogens,  NTP 81-43, Dec.  1981,  pp.
      236-237.

9.    Blum,  A., Ames, B.N., Science 195, 1977,  17.

                                  8321A -  23                        Revision 1
                                                                  January 1995

-------
10.  Zweidinger,  R.A.,  Cooper,  S.D.,  Pellazari,  E.D.,  Measurements of Organic
     Pollutants  in Water and Wastewater, ASTM 686.

11.  Cremlyn,  R.,  Pesticides:  Preparation  and mode of Action,  John  Wiley and
     Sons, Chichester,  1978, p. 142.

12.  Cotterill,  E.G., Byast, T.H., "HPLC of Pesticide Residues in Environmental
     Samples",  In Liquid Chromatography in  Environmental  Analysis,  Laurence,
     J.F., Ed.,  Humana  Press, Clifton,  NJ, 1984.

13.  Voyksner, R.D., "Thermospray HPLC/MS  for Monitoring the Environment",  In
     Applications of New Mass  Spectrometry Techniques  in Pesticide Chemistry;
     Rosen, J.D., Ed.,  John Wiley and Sons:  New York, 1987.

14.  Yinon, J.,  Jones,  T.L., Betowski,  L.D., Rap.  Comm.  Mass Spectrom., 1989,
     3, 38.

15.  Shore,  F.L., Amick,  E.N.,  Pan,  S.T.,  Gurka,  D.F.,   "Single  Laboratory
     Validation  of EPA  Method  8150  for  the Analysis of Chlorinated Herbicides
     in  Hazardous  Waste",  EPA/600/4-85/060,  U.S.  Environmental  Protection
     Agency, Las Vegas, NV, 1985.

16.  "Development and  Evaluations of  an LC/MS/MS Protocol", EPA/600/X-86/328,
     Dec. 1986.

17.  "An  LC/MS Performance Evaluation  Study of  Organophosphorus Pesticides",
     EPA/600/X-89/006,  Jan. 1989.

18.  "A   Performance    Evaluation   Study  of  a   Liquid  Chromatography/Mass
     Spectrometry   Method     for     Tris-(2,3-Dibromopropyl)     Phosphate",
     EPA/600/X-89/135,  June 1989.

19.  "Liquid   Chromatography/Mass  Spectrometry   Performance   Evaluation  of
     Chlorinated Phenoxyacid  Herbicides and Their Esters", EPA/600/X-89/176,
     July 1989.

20.  "An  Inter!aboratory Comparison of an SW-846  Method for the Analysis of the
     Chlorinated Phenoxyacid Herbicides by  LC/MS",  EPA/600/X-90/133, June 1990.

21.  Somasundaram, L.,  and J.R. Coates,  Ed.,  "Pesticide Transformation Products
     Fate and  Significance in  the Environment",   ACS  Symposium  Series 459,
     Ch.  13, 1991.

22.  Single-Laboratory  Evaluation of  Carbamates, APPL, Inc., Fresno,  CA.

23.  "Interlaboratory Calibration Study of  a  Thermospray-Liquid Chromatography/
     Mass Spectrometry (TS-LC/MS) Method  for Selected Carbamate Pesticides",
     EPA/600/X-92/102,  August  1992.
                                  8321A - 24                        Revision 1
                                                                  January 1995

-------
                              TABLE  1
            RECOMMENDED HPLC CHROMATOGRAPHIC  CONDITIONS
Initial
Mobile Initial
Phase Time
(%) (min)
Analytes:
Orqanophosphorus Compounds
50/50 0
(water/methanol)
Azo Dyes (e.g. Disperse Red 1)
50/50 0
(water/CH3CN)
Tris-(2,3-dibromopropyl ) phosphate
50/50 0
(water/methanol)
Final
Gradient Mobile
(linear) Phase
(min) (%)
10 100
(methanol )
5 100
(CH3CN)
10 100
(methanol)
Final
Time
(min)
5
5
5
Chlorinated phenoxyacid compounds

75/25                   2
(A/methanol)

40/60                   3
(A/methanol)
                                    15       40/60
                                          (A/methanol)

                                             75/25
           10
                                          (A/methanol)

Where A = 0.1 M ammonium acetate (1% acetic acid)

Carbamates

Option A:
Time
(mini

  0
 30
 35
 40
 45
                     Mobile phase A
                         (percent)

                             95
                             20
                              0
                             95
                             95
Mobile phase B
     (percent)

         5
        80
       100
         5
         5
Where A = 5 mM ammonium acetate with 0.1 M acetic acid
and   B = methanol
With optional post-column addition of 0.5 M ammonium acetate.
                            8321A - 25
                                                              Revision 1
                                                            January 1995

-------
Carbamates (continued)

Option B:
TABLE 1  (continued)


                                                      €
                     Mobile phase A          Mobile phase B
                         (percent)                (percent)

  0                          95                       5
 30                           0                     100
 35                           0                     100
 40                          95                       5
 45                          95                       5

Where A = water with 0.1 M ammonium acetate with 1% acetic acid
      B = methanol with 0.1 M ammonium acetate with 1% acetic acid
with optional post-column addition of 0.1 M ammonium acetate.
                            8321A  - 26                        Revision  1
                                                            January  1995

-------
                                    TABLE  2
              COMPOUNDS AMENABLE  TO  THERMOSPRAY MASS SPECTROMETRY
    Disperse Azo Dyes            Alkaloids
    Methine Dyes                 Aromatic ureas
    Arylmethane Dyes             Amides
    Coumarin Dyes                Amines
    Anthraquinone Dyes           Amino acids
    Xanthene Dyes                Organophosphorus Compounds
    Flame retardants             Chlorinated Phenoxyacid Compounds
    Carbamates
                                    TABLE  3
                 LIMITS OF DETECTION AND METHOD SENSITIVITIES
                        FOR DISPERSE RED 1 AND  CAFFEINE
Compound
Disperse Red 1


Caffeine


Mode
SRM
Single Quad
CAD
SRM
Single Quad
CAD
LOD
P9
180
600
2,000
45
84
240
EQL(7s)
P9
420
1400
4700
115
200
560
EQL(lOs)
pg
600
2000
6700
150
280
800
EQL = Estimated Quantitation Limit

Data from Reference 16.
                                  8321A - 27                        Revision 1
                                                                  January 1995

-------
                                    TABLE  4
            PRECISION  AND  ACCURACY  COMPARISONS OF MS AND MS/MS WITH
       HPLC/UV  FOR  ORGANIC-FREE  REAGENT  WATER SPIKED WITH DISPERSE RED  1
                                       Percent Recovery
Sample
HPLC/UV
MS
CAD
SRM
Spike 1
Spike 2
RPD
82.2 ± 0.2
87.4 + 0.6
6.1%
92.5 + 3.7
90.2 ± 4.7
2.5%
87.6 ± 4.6
90.4 ± 9.9
3.2%
95.5 + 17
90.0 ± 5.
5.9%
.1
9

Data from Reference 16.
                                    TABLE  5
           PRECISION AND ACCURACY COMPARISONS OF MS AND MS/MS WITH
          HPLC/UV FOR MUNICIPAL  WASTEWATER SPIKED WITH  DISPERSE  RED  1

Sample
Spike 1
Spike 2
RPD

HPLC/UV
93.4 ± 0
96.2 ± 0
3.0%
Percent Recovery
MS
.3 102.0 ± 31
.1 79.7 + 15
25%

CAD
82.7 + 13
83.7 ± 5.2
1.2%
Data from Reference 16.
                                  8321A - 28
                                                 Revision 1
                                               January 1995

-------
                                    TABLE  6
         RESULTS FROM ANALYSES OF ACTIVATED SLUDGE PROCESS WASTEWATER
Sample
5 mg/L Spiking
Concentration
1
1-D
2
3
RPD
0 mg/L Spiking
Concentration
1
1-D
2
3
RPD
Recovery
HPLC/UV

0.721 ± 0.003
0.731 ± 0.021
0.279 ± 0.000
0.482 ± 0.001
1.3%

0.000
0.000
0.000
0.000
--
of Disperse Red 1
MS

0.664 + 0.030
0.600 + 0.068
0.253 + 0.052
0.449 ± 0.016
10.1%

0.005 + 0.0007
0.006 ± 0.001
0.002 ± 0.0003
0.003 + 0.0004
18.2%
(mq/L)
CAD

0.796 + 0.008
0.768 + 0.093
0.301 + 0.042
0.510 + 0.091
3.6%

<0.001
<0.001
<0.001
<0.001
--
Data from Reference 16.
                                  8321A - 29
  Revision 1
January 1995

-------
                   TABLE 7
CALIBRATION MASSES AND % RELATIVE ABUNDANCES
                  OF  PEG 400
Mass
18.0
35.06
36.04
50.06
77.04
168.12
212.14
256.17
300.20
344.22
388.25
432.28
476.30
520.33
564.35
608.38
652.41
653.41
696.43
697.44
% Relative
Abundances8
32.3
13.5
40.5
94.6
27.0
5.4
10.3
17.6
27.0
45.9
64.9
100
94.6
81.1
67.6
32.4
16.2
4.1
8.1
2.7
   Intensity is normalized to mass 432.
                  8321A -  30                        Revision 1
                                                  January 1995

-------
                   TABLE 8
CALIBRATION MASSES AND % RELATIVE ABUNDANCES
                 OF PEG 600
Mass
18.0
36.04
50.06
77.04
168.12
212.14
256.17
300.20
344.22
388.25
432.28
476.30
520.33
564.35
608.38
652.41
653.41
696.43
% Relative
Abundances"
4.7
11.4
64.9
17.5
9.3
43.9
56.1
22.8
28.1
38.6
54.4
64.9
86.0
100
63.2
17.5
5.6
1.8
         Intensity  is  normalized  to mass  564.
                 8321A -  31                         Revision 1
                                                 January 1995

-------
                                   TABLE 9
                 RETENTION TIMES AND THERMOSPRAY MASS SPECTRA
                        OF ORGANOPHOSPHORUS COMPOUNDS
Compound
Monocrotophos
Trichlorfon
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
Retention Time
(minutes)
1:09
1:22
1:28
4:40
9:16
9:52
10:52
13:30
13:55
18:51
Mass Spectra
(% Relative Abundance)"
241 (100), 224 (14)
274 (100), 257 (19), 238 (19)
230 (100), 247 (20)
238 (100), 221 (40)
398 (100), 381 (23), 238 (5),
221 (2)
326 (10), 309 (100)
281 (100), 264 (8), 251 (21),
234 (48)
278 (4), 261 (100)
292 (10), 275 (100)
315 (100), 299 (15)
  a  For molecules containing Cl,  Br and S, only the base peak of the isotopic
  cluster is listed.

Data from Reference 17.
                                  8321A - 32
  Revision 1
January 1995

-------
                                   TABLE 10
                     PRECISION AND LIMITS OF DETECTION FOR
                      ORGANOPHOSPHORUS COMPOUND STANDARDS
Compound
Dichlorvos
Dimethoate
Phorate
Disulfoton
Fensulfothion
Naled
Merphos
Methyl
parathion
Ion
238
230
261
275
309
398
299
281
Standard
Quantitation
Concentration
(ng/ML)
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
%RSD
16
13
5.7
4.2
2.2
4.2
13
7.3
0.84
14
7.1
4.0
2.2
14
6.7
3.0
4.1
9.2
9.8
2.5
9.5
9.6
5.2
6.3
5.5
17
3.9
5.3
7.1
4.8
1.5
MDL (ng)
4
2
2
1
0.4
0.2
1
30
Data from Reference 17.
                                  8321A - 33
  Revision 1
January 1995

-------
                                   TABLE 11
    SINGLE OPERATOR ACCURACY AND PRECISION FOR LOW CONCENTRATION DRINKING
     WATER (A),  LOW CONCENTRATION SOIL (B), MEDIUM CONCENTRATION DRINKING
                 WATER (C),  MEDIUM CONCENTRATION  SEDIMENT (D)
Average
Recovery
Compound (%)
A
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
B
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
C
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
D
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos

70
40
0.5
112
50
16
3.5
237

16
ND
ND
45
ND
78
36
118

52
146
4
65
85
10
2
101

74
166
ND
72
84
58
56
78
Standard
Deviation

7.7
12
1.0
3.3
28
35
8
25

4


5

15
7
19

4
29
3
7
24
15
1
13

8.5
25

8.6
9
6
5
4
Spike
Amount
uq/L
5
5
5
5
10
5
5
5
ttg/q
50
50
50
50
100
50
50
50
uq/L
50
50
50
50
100
50
50
50
mq/kq
2
2
2
2
3
2
2
2
Range of
Recovery
(%)

85 -
64 -
2 -
119 -
105 -
86 -
19 -
287 -

24 -


56 -

109 -
49 -
155 -

61 -
204 -
9 -
79 -
133 -
41 -
4 -
126 -

91 -
216 -

90 -
102 -
70 -
66 -
86 -

54
14
0
106
0
0
0
187

7


34

48
22
81

43
89
0
51
37
0
0
75

57
115

55
66
46
47
70
Number
of
Analyses

15
15
15
15
15
15
15
15

15
15
15
15
15
15
15
15

12
12
12
12
12
12
12
12

15
15
15
15
15
15
15
12
Data from Reference 17.
                                  8321A - 34
  Revision 1
January 1995

-------
                                   TABLE 12
          SINGLE OPERATOR ACCURACY AND PRECISION FOR MUNICIPAL WASTE
           WATER (A), DRINKING WATER (B), CHEMICAL SLUDGE WASTE (C)
Compound
Tris-BP (A)
(B)
(C)
Average
Recovery
(%)
25
40
63
Spike
Standard Amount
Deviation (ng/juL)
8.0 2
5.0 2
11 100
Range
of %
Recovery
41 - 9.0
50 - 30
84 - 42
Number of
Analyses
15
12
8
Data from Reference 18.

Concentration
(ng/juL)
50
100
150
200


SINGLE
Average
Area
2675
5091
7674
8379
LOD
(ngM)
33
TABLE 13
OPERATOR EQL TABLE FOR
Standard 3*Std
Deviation Dev.
782 2347
558
2090
2030
Lower
EQL
(ng//*L)
113
TRIS-BP
7*Std
Dev.
5476
Upper
EQL
(ng/W
172

10*Std
Dev.
7823


EQL = Estimated Quantitation Limit

Data from Reference 18.
                                  8321A - 35
  Revision 1
January 1995

-------
                                   TABLE 14
          LIMITS OF DETECTION IN THE POSITIVE AND NEGATIVE ION MODES
          FOR THE CHLORINATED PHENOXYACID HERBICIDES AND FOUR ESTERS
Compound
Dalapon
Dicamba
2,4-D
MCPA
Dichlorprop
MCPP
2,4,5-T
2,4,5-TP (Silvex)
Dinoseb
2,4-DB
2,4-D,Butoxy
ethanol ester
2,4,5-T,Butoxy
ethanol ester
2,4,5-T, Butyl
ester
2,4-D,ethyl-
hexyl ester
Positive Mode
Quantitation
Ion
Not detected
238 (M+NH4) +
238 (M+NH4)+
218 (M+NH4)+
252 (M+NH4)+
232 (M+NH4)+
272 (M+NH4)+
286 (M+NH4)+
228 (M+NH4-NO)+
266 (M+NH4) +
321 (M+Hr

372 (M+NH4) +

328 (M+NH4)+

350 (M+NH4)+

LOD
(ng)

13
2.9
120
2.7
5.0
170
160
24
3.4
1.4

0.6

8.6

1.2

Negative Mode
Quantitation
Ion
141 (M'H)'
184 (M'HCl)-
184 (M'HCl)-
199 (Ml)'
235 (Ml)'
213 (M'l)-
218 (M'HCl)-
269 (M'l)-
240 (M)'
247 (M'l)-
185 (MXeH^O,)-

195 (M-C8H1503)-

195 (M-CeH^O^-

161 (M-C10H1903)-

LOD
(ng)
11
3.0
50
28
25
12
6.5
43
19
110








Data from Reference 19.
                                  8321A - 36
  Revision 1
January 1995

-------
                                  TABLE  15
              SINGLE LABORATORY OPERATOR ACCURACY AND PRECISION
                  FOR THE  CHLORINATED  PHENOXYACID HERBICIDES
Compound
(a)
Average
Recovery(%)
Standard
Deviation
Spike
Amount
Range of
Recovery
Number
of
Analyses
               LOW LEVEL DRINKING WATER
  Dicamba
   2,4-D
   MCPA
   MCPP
Dichlorprop
  2,4,5-T
  Silvex
  2,4-DB
  Dinoseb
  Dalapon
2,4-D,ester
  Dicamba
   2,4-D
   MCPA
   MCPP
Dichlorprop
  2,4,5-T
  Silvex
  2,4-DB
  Dinoseb
  Dalapon
2,4-D,ester
  Dicamba
   2,4-D
   MCPA
   MCPP
Dichlorprop
  2,4,5-T
  Silvex
  2,4-DB
  Dinoseb
  Dalapon
2,4-D,ester
63
26
60
78
43
72
62
29
73
ND
73
HIGH LEVEL DRINKING
54
60
67
66
66
61
74
83
91
43
97
LOW LEVEL SAND
117
147
167
142
ND
134
121
199
76
ND
180
22
13
23
21
18
31
14
24
11
ND
17
WATER
30
35
41
33
33
23
35
25
10
9.6
19

26
23
79
39
ND
27
23
86
74
ND
58
 5
 5
 5
 5
 5
 5
 5
 5
 5
 5
 5
50
50
50
50
50
50
50
50
50
50
50
 M9/L

 86 - 33
 37 -  0
 92 - 37
116 - 54
 61 -  0
138 - 43
 88 - 46
 62 -  0
 85 - 49
     ND
104 - 48
                                                          M9/L
103
119
128
122
116
 99
132
120
102
 56
130
26
35
32
35
27
44
45
52
76
31
76
            M9/9

         147 -  82
         180 -118
         280 -  78
         192 -  81
              ND
         171 -  99
         154 -  85
         245 -  0
         210 -  6
              ND
         239 -  59
              9
              9
              9
              9
              9
              9
              9
              9
              9
              9
              9
9
9
9
9
9
9
9
9
9
6
9
                    10
                    10
                    10
                    10
                    10
                    10
                    10
                    10
                    10
                    10
                     7
 (a)All  recoveries  are  in  negative  ionization mode, except  for  2,4-D,ester.
                             ND = Not Detected.
                                 8321A - 37
                    Revision 1
                  January 1995

-------
                            TABLE 15  (continued)
              SINGLE LABORATORY OPERATOR ACCURACY  AND PRECISION
                 FOR THE CHLORINATED  PHENOXYACID HERBICIDES


Compound
(a)
Average
Recovery(%)

Standard
Deviation

Spike
Amount
Range of
Recovery
(%)
Number
of
Analyses
  Dicamba
   2,4-D
   MCPA
   MCPP
Dichlorprop
  2,4,5-T
  Silvex
  2,4-DB
  Dinoseb
  Dalapon
2,4-D,ester
  Dicamba
   2,4-D
   MCPA
   MCPP
Dichlorprop
  2,4,5-T
  Silvex
  2,4-DB
  Dinoseb
  Dalapon
2,4-D,ester
  Dicamba
   2,4-D
   MCPA
   MCPP
Dichlorprop
  2,4,5-T
   Silvex
   2,4-DB
  Dinoseb
  Dalapon
2,4-D,ester
               HIGH LEVEL SAND
153
218
143
158
92
160
176
145
114
287
20
LOW LEVEL MUNICIPAL
83
ND
ND
ND
ND
27
68
ND
44
ND
29
HIGH LEVEL MUNICIPAL
66
8.7
3.2
10
ND
2.9
6.0
ND
16
ND
1.9
33
27
30
34
37
29
34
22
28
86
3.6
ASH
22
ND
ND
ND
ND
25
38
ND
13
ND
23
ASH
21
4.8
4.8
4.3
ND
1.2
3.1
ND
6.8
ND
1.7
209 -119
276 -187
205 -111
226 -115
161 - 51
204 -131
225 -141
192 -110
140 - 65
418 -166
 25 - 17

 M9/9

104 - 48
     ND
     ND
     ND
     ND
 60 -  0
128 - 22
     ND
 65 - 26
     ND
 53 -  0

 M9/9
 96
 21
 10
 16
     41
      5
      0
     4.7
    ND
3.6-  0
12 - 2.8
    ND
23 -  0
    ND
6.7-  0
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    7
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    9
                    6
9
9
9
9
9
9
9
9
9
9
6
  ''All  recoveries are in negative ionization mode,  except  for 2,4-D,ester.
                              ND  =  Not  Detected.
                                  8321A  -  38
           Revision 1
         January 1995

-------
                                   TABLE 16
                 MULTI-LABORATORY ACCURACY AND PRECISION DATA
                  FOR THE CHLORINATED PHENOXYACID HERBICIDES
        Compounds
Spiking
Concentration
                                              Mean
                 %  Relative
(% Recovery)" Standard Deviation

2,4,5-T
2,4,5-T,butoxy
2,4-D
2,4-DB
Dalapon
Dicamba
Dichlorprop
Dinoseb
MCPA
MCPP
Silvex
500 mq/L
90
90
86
95
83
77
84
78
89
86
96

23
29
17
22
13
25
20
15
11
12
27
                                              50 mq/L

            2,4,5-T                                  62
        2,4,5-T,butoxy                              85
             2,4-D                                   64
            2,4-DB                                 104
            Dalapon                                 121
            Dicamba                                  90
          Dichlorprop                                96
            Dinoseb                                  86
             MCPA                                   96
             MCPP                                   76
            Silvex                                  65

                                              5 mq/L

            2,4,5-T                                  90
        2,4,5-T,butoxy                              99
             2,4-D                                  103
            2,4-DB                                  96
            Dalapon                                 150
            Dicamba                                 105
          Dichlorprop                               102
            Dinoseb                                 108
             MCPA                                   94
             MCPP                                   98
            Silvex                                  87

"  Mean  of duplicate data from 3  laboratories.
b  % RSD of duplicate  data from 3 laboratories.

Data from Reference 20.
                                  8321A - 39
                                                  68
                                                   9
                                                  80
                                                  28
                                                  99
                                                  23
                                                  15
                                                  57
                                                  20
                                                  74
                                                  71
                                                  28
                                                  17
                                                  31
                                                  21
                                                   4
                                                  12
                                                  22
                                                  30
                                                  18
                                                  15
                                                  15
                                          Revision 1
                                        January 1995

-------
                                   TABLE 17
           COMPARISON OF LODs:  METHOD 8151 vs.  METHOD 8321
lonization
Compound
     Method 8151
Aqueous Samples
     GC/ECD
 EDLa (Mg/L)
  Method 8321
Aqueous Samples
      HPLC/MS/TS
   LOD (Mg/L)
                                                                      Mode
Dalapon
Dicamba
2,4-D
MCPA
Dichlorprop
MCPP
2,4,5-T
2,4,5-TP (Silvex)
2,4-DB
Dinoseb
1.3
0.081
0.2
0.056b
0".26
0.09
0.08
0.075
0.8
0.19
1.1
0.3
0.29
2.8
0.27
0.50
0.65
4.3
0.34
1.9
(-)
(-)
(+)
(-)
(+)
(+)
(-)
(-)
(+)
(-)
   EDL = estimated detection limit; defined as either the MDL,  or a
   concentration of analyte in a sample yielding a peak in the  final  extract
   with signal-to-noise ratio of approximately 5,  whichever value is  higher.
   40 CFR Part 136, Appendix B (49 FR 43234).
   capillary column.
                      Chromatography using wide-bore
                                  8321A - 40
                                           Revision 1
                                         January 1995

-------
                                 TABLE 18
          SINGLE-LABORATORY METHOD DETECTION LIMIT DETERMINATION
                      AND  PRECISION RESULTS  - WATERC
Analyte
Aldicarb sulfoxide"
Aldicarb sulfone
Oxamyl
Methomyl
3-Hydroxycarbofurana
Fenuron
Benomyl/Carbendazim
Aldicarb
Aminocarb
Carbofuran
Propoxur
Monuron
Bromacil
Tebuthiuron
Carbaryl
Fluometuron
Propham
Propachlor
Diuron
Siduron
Methiocarb
Barban
Linuron
Chloropropham
Mexacarbate
Chloroxuron
Neburon
Average %
Recovery
7.5
88.4
60.7
117
37.4
104
67.3
93.7
117
94.2
106
95.6
86.4
106
85.1
89.1
84.2
98.5
95.6
105
92.4
90.5
97.7
89.1
80.0
109
92.5
Standard
Deviation
0.27
0.44
0.10
0.49
0.25
0.20
0.13
0.46
0.53
0.17
0.32
0.24
0.12
0.17
0.29
0.19
0.15
0.16
0.14
0.27
0.16
0.79
0.19
0.68
1.41
0.32
0.14
%RSD
72.4
50.3
16.6
41.5
65.4
19.3
19.7
49.6
44.9
17.7
30.4
25.6
14.1
16.1
34.1
21.7
17.3
16.0
14.7
25.9
17.5
17.4
19.5
15.2
35.1
29.2
14.9
MDLb
M9/L
0.8
1.3
0.3
1.5
0.8
0.6
0.4
1.4
1.6
0.5
1.0
0.7
0.4
0.5
0.9
0.6
0.4
0.5
0.4
0.8
0.5
2.4
0.6
2.0
4.2
1.0
0.4
- Values generated from  internal response factor calculations.

- Method detection limit determinations are based on twenty water
  extractions.  Aldicarb sulfoxide, Barban, Chloropropham, and Mexacarbate
  spike levels were  at 5 /ug/L.   All other analytes  were  spiked at  1  M9/L.
  The method detection limit was determined by multiplying the standard
  deviation by 3.  Quantitation was done using average linear regression
  values, unless otherwise  indicated.

- Data from Reference 22.
                                8321A - 41                        Revision 1
                                                                January 1995

-------
                                 TABLE 19
         SINGLE-LABORATORY METHOD QUANTITATION  LIMIT  DETERMINATION
                       AND PRECISION RESULTS -  SOILb
Analyte
Aldicarb sulfoxide
Aldicarb sulfone
Oxamyl
Methomyl
3-Hydroxycarbofuran
Fenuron
Benomyl/Carbendazim
Aldicarb
Aminocarb
Carbofuran
Propoxur
Monuron
Bromacil
Tebuthiuron
Carbaryl
Fluometuron
Propham
Propachlor
Diuron
Siduron
Methiocarb
Barban
Linuron
Chloropropham
Mexacarbate
Chloroxuron
Neburon
Average %
Recovery
66.9
118
89.6
86.8
103
91.2
68.0
72.0
84.4
102
95.2
107
99.6
96.8
99.6
92.8
100
114
101
107
124
108
113
104
62.2
97.6
110
Standard
Deviation
0.0492
0.0076
0.0049
0.0051
0.0116
0.0049
0.0082
0.0056
0.0082
0.0083
0.0091
0.0077
0.0069
0.0071
0.0054
0.0035
0.0039
0.0037
0.0060
0.0063
0.0054
0.0333
0.0037
0.0217
0.0119
0.0031
0.0044
%RSD
58.9
25.7
21.9
23.6
45.0
21.6
47.0
30.1
38.7
32.7
38.2
28.8
27.5
29.5
21.7
15.1
15.7
13.0
23.8
23.7
17.5
24.8
13.0
16.6
15.3
12.6
16.0
MDL"
M9/9
0.15
0.023
0.015
0.015
0.035
0.015
0.025
0.017
0.025
0.025
0.027
0.023
0.021
0.021
0.016
0.011
0.012
0.011
0.018
0.019
0.016
0.10
0.011
0.065
0.036
0.009
0.011
- Method detection  limit determinations  are based on twenty soil
  extractions.  Aldicarb sulfoxide,  Barban, Chloropropham, and Mexacarbate
  spike levels were at  0.125 M9/9-   All  other  analytes  were spiked  at  0.025
  M9/9-  The method detection  limit  was  determined  by multiplying the
  standard deviation by 3.  Quantitation was done using average linear
  regression values,  unless otherwise  indicated.

- Data from Reference 22.
                                8321A - 42                        Revision 1
                                                                January 1995

-------
                                 TABLE 20
             SINGLE-LABORATORY EVALUATION OF AVERAGE RECOVERY
                        AND PRECISION DATA - WATER0
Analyte
Aldicarb sulfoxide
Aldicarb sulfone
Oxamyl8
Methomyl
3 -Hydroxycarbof uran"
Fenuron
Benomyl/Carbendazim
Aldicarb
Aminocarb
Carbofuran
Propoxur
Monuron
Bromacil
Tebuthiuron
Carbaryl
Fluometuron
Propham
Propachlor
Diuron
Siduron
Methiocarb
Barban
Linuron
Chloropropham
Mexacarbate
Chloroxuron
Neburon
Average %
Recovery6
7.6
56.0
38.9
52.0
22.2
72.5
47.3
81.0
109
85.5
79.1
91.8
87.6
87.1
82.1
84.4
80.7
84.3
90.8
88.0
93.3
88.1
87.1
94.9
79.8
106
85.3
Standard
Deviation
2.8
27.1
17.9
19.6
9.3
22.0
14.7
13.7
38.3
10.0
13.7
11.3
12.1
9.0
13.5
8.3
13.8
10.0
14.1
9.5
12.8
11.2
16.8
15.3
12.9
24.9
12.6
%RSD
37.0
48.5
45.9
37.7
41.7
30.3
31.0
16.9
35.1
11.7
17.3
12.3
13.8
10,3
16.5
9.8
17.1
11.9
15.6
10.8
13,8
12.7
19.3
16.1
16.2
23.5
14.8
- Values generated  from  internal  response  factor calculations.

- Nine  spikes were  performed  at three  concentrations.  The  concentrations
  for Aldicarb  sulfoxide,  Barban,  Chloropropham, and Mexacarbate  spike
  levels were at  25 /zg/L,  50  fj.g/1, and 100 /ig/L.   All  other analyte
  concentrations  were  5  M9/U  10  |ug/L, and 20 M9/L.  One injection was
  disregarded as  an outlier.   The  total number of  spikes  analyzed was 26.
  Quantitation  was  done  using  average  linear  regression values, unless
  otherwise  indicated.

- Data from Reference  22.
                                8321A - 43                        Revision 1
                                                                January 1995

-------
                                 TABLE 21
             SINGLE-LABORATORY EVALUATION OF AVERAGE RECOVERY
                        AND PRECISION DATA - SOIL6
Analyte
Aldicarb sulfoxide
Aldicarb sulfone
Oxamyl
Methomyl
3-Hydroxycarbofuran
Fenuron
Benomyl/Carbendazim
Aldicarb
Aminocarb
Carbofuran
Propoxur
Monuron
Bromacil
Tebuthiuron
Carbaryl
Fluometuron
Propham
Propachlor
Diuron
Siduron
Methiocarb
Barban
Linuron
Chloropropham
Mexacarbate
Chloroxuron
Neburon
Average %
Recovery8
66.9
162
78.9
84.9
105
91.9
95.6
97.9
133
109
104
101
100
104
102
' 94.5
92.8
94.6
107
100
107
92.3
104
105
77.2
121
92.1
Standard
Deviation
31.3
51.4
46.1
25.8
36.3
16.7
18.2
17.0
44.7
14.4
16.5
12.4
9.0
11.9
15.5
15.7
12.0
10.3
17.4
12.0
14.2
15.6
13.6
9.3
9.8
27.3
16.5
%RSD
46.7
31.7
58.5
30.4
34.5
18.1
19.0
17.4
33.6
13.2
15.9
12.3
9.0
11.5
15.2
16.7
12.9
10.9
16.2
12.0
13.2
16.9
13.1
8.9
12.7
22.5
17.9
- Nine  spikes were  performed  at  three concentrations.  The concentrations
  for Aldicarb  sulfoxide,  Barban, Chloropropham, and Mexacarbate  spike
  levels were at  0.625  jug/g,  1.25 jug/g,  and 2.5 M9/9-   All  other analyte
  concentrations  were 0.125 ng/g, 0.25 jug/9»  and 0.50  jug/g.   One injection
  was disregarded as an outlier.  The total number  of  spikes  analyzed was
  26.   Quantitation was done  using  average  linear regression  values.

- Data from Reference 22.
                                8321A  - 44
  Revision 1
January 1995

-------
                                   TABLE 22
                MULTI-LABORATORY EVALUATION OF METHOD ACCURACY
                            (AFTER OUTLIER REMOVAL)"
                                       Percent Recovery
Analyte
High-Concentration
     Samples8
Medium-Concentration
       Samplesb
Low-Concentration
    Samples0
Aldicarb
Bendiocarb
Carbaryl
Carbendazim
Carbofuran
Diuron
Linuron
Methomyl
Oxamyl
98.7
81.4
92.0
125
87.8
79.9
84.8
93.3
83.8
110
95.0
108
138
92.3
98.8
93.0
90.8
88.0
52.0
52.0
62.0
128
72.0
66.0
82.0
90.0
98.0
  -  Three replicates per laboratory; eight to nine laboratories (per Table 26
    of Reference 23).  The true concentration is 90 mg/L per compound, except
    Carbendazim at 22.5 mg/L.

  -  Two replicates per laboratory; eight to nine laboratories (per Table 26
    of Reference 23).  The true concentration is 40 mg/L per compound except
    Carbendazim at 10 mg/L.

  -  Three replicates per laboratory; eight to nine laboratories (per Table 26
    of Reference 23).  The true concentration is 5 mg/L per compound, except
    Carbendazim at 1.25 mg/L.

  -  Data  from Reference  23.
                                  8321A - 45
                                                 Revision  1
                                               January 1995

-------
   o
   I-H
   00
   o
   LU
   o
   Q,
   Ofi
co

   O
   QQ
             CU
             O
             C
             o
            o


             o
 re
 i-


 cu
 u

 o
o

 £


•5
 cu
             ra
             s_

             c
             O)
             o
             c
             o
            o
             en
                  oo
                  O£
                  a
                  oo
                  a:
                   en
  a:
Q
00
a
oo
QL
                    CC

                   C/)
                   cn

                  
00"—i in oo •—i r- •—ICMCO    en    vo    ... ra
enior»-ioenr»»incnen    r-.    •—i    _~

                                             o
•—i co co •—i CM CM r>« co r>.    i>~    •—i    &

co>—ico^oiniomincn    o    »-~    a
CO CM CM CM CM rt •—1 •—I       CM


                                          °§
VOlOCOi-HCOVO-H,-HVO                00 -<

CMi—icMi—ICOCMCM^-«*-                	
                                          T3
                                          C
                                          re

enior-^.^cnLovD^in
                                          o
OOOOOOOOO                00
                                          Qi Q)


icvOi—iiovocO"—imen                ^

CMCMcO'—icoco«cJ"^i-«i-                'eo<'u

                                          ^ (/)

                                          2 2
                                          o
                                          CO  ^
inr^.^cMcoooencMi-H    r>-    CM    Q.
                                          V)
oocoio^i-^»-cnini—icn    co    •—i    cu.^-
co'J'covo^-cMco^-in    ^-    ^H    i_i	

                                           -^
incoo^i-voininoo^-    CM
                                             o
         ocovoor>»o    •—i    ^-    •'-s
                                             •o
                                             o

eoior^cncooo^-ooo                "0°"

      in oo 10 t—i co in o                ^
            1—I I—I r-l I—I CM                g


                                          al
   ioo«3-oujcnoor^                co —

   VOCO>—linCMCOCMCO


                                          re '.r-
   o^HooenmcMcocM

   oococotocnr>«voin
   co^t-i-Hcococococo                •'-ns
                                          r— Q)
                                          ••- a.
                                          •2 w
com-—«^j-mco^ooi—i    en    <—>    4j*~

oocoi—i^t-^rvDCMincn    co    i^.    JJJ4-
com^-in^-co«d-co^-    ^-          Q.I£

                                          co
                                          i- •-
      cMcncMCMcncn«i-    in    o

cMi—i^-cni—IOOOCMIO    10    «*    o.2
i-H CM i-H i-H CM i-H i-H i-H i-H    	•          •

                                             ra

   coococMi—iin^-o                c. "~
                                          o co
         ininvocMcni-.                •<- -a
COCOCO-—ICOCMCOCMCO
                                             •o

•

                                          ^5 «
oocooo^Hoencooin                -UQ,

oococMOOcn'—iio-«t-in                — >
OOf^OOCMI^.t^.h^OOt^-                I * +J

                                          M *J
                                          CO a) ^^
                                          .C S-  CT>
                                          -ME
                                             cn

                                          2! =  £
                                          ^O ^3  nj
                                           — C   _
                   ^    -Ec=
                   S-    N re
                Xi re i— re S*.       i—             >
                S- U >>~O 3    C >>       CU    CU
                «oi-c<4-eoEi—    cn   Q
                U'r-»OCUOOi-O>>       4->
                ctCQOOOQ—ISIO    «C   OO
                                                                                                           •-< in
                                                                                                             en
                                                                                                           c en
                                                                                                           O f-H
                                                                                                                       0) 3
                                                                                                                       o: c
                                                                                                                       CM

                                                                                                                       CO
                                                                                                                       00
S
                                                    =: co ^
                                                    b- C  C
                                                    fl ^ «

                                                     .. o   ^
                                                    y) (j to
                                                                                  co
                                                                                  CM
                                                                                  o
                                                                                  s_
                                                                                  cu

                                                                                  cu
                                                                                  oc:
                                                                                  o
                                                                                  S-
                                                       08


                                                      o

-------
                                    FIGURE 1
                SCHEMATIC OF THE THERMOSPRAY PROBE AND ION SOURCE
                                    Flange
   To
   Trap
    ft —
Mechanical
  Pump
               I
 Source
Mounting
  Plate
               I
•lock
             Ion Sampling
                 Cone
Ion*
electron  Vaporiser
 team
I
                                                             — LC
             Vapor
           Tomporaturt  |

            T4       Hock
                  Temperature
                   T.
                                                     Coupling
                                    8321A - 47
                                                   Revision  1
                                                January  1995

-------
            FIGURE 2
THERMOSPRAY SOURCE WITH WIRE-REPELLER
   (High  sensitivity configuration)
                                                        i
      CERAMIC INSULATOR
      WIRE REPELLER
            8321A - 48
  Revision 1
January 1995

-------
              FIGURE 3
         THERMOSPRAY SOURCE WITH WIRE-REPELLER
            (CAD configuration)
rxx
          CERAMIC INSULATOR

          WIRE REPELLER
             8321A - 49
 Revision 1
January 1995

-------
                                                   METHOD 8321
                           SOLVENT EXTRACTABLE  NON-VOLATILE COMPOUNDS BY
           HIGH PERFORMANCE LIQUID CHROMATOGRAPHY/THERMQSPRAY/MASS SPECTROHETRY
                              (HPLC/TS/MS)  OR ULTRAVIOLET (UV)  DETECTION
                                                l« aample
                                              analyzed for
                                                Trie-BP7
                      7.1 Prepare aample
                        for extreetion.
                             U
                         aample higK
                        concentration
                           wait*?
 Uae dilution
Method 3580.
                                     Chlorinated Phenoxyacid
                                          Compound!
                                       7.1.2 U«. modified
                                         Method 8151.
                    Carbemate
                    Peeticidea
 uee extraction
Method 3540 or
 Method 3550.
                 Solid /aemple aolid
                        or equeoue?
                         Ua* •xtraction
                        Method 3610 or
                         Mothod 3620.
                         7.2 Exchange
                        • xtraetion •olvant
                         to mathanol or
                        aeatomtrila dunng
                        K-D procadurae.
                          7 3 3«t HPLC
                        Chroma tographie
                          condition*.
                    7.1.1 Prepare aample
                        for Tria-BP
                      microextrection.
  7.1.1.1.1 Spike
the (elected (ample
   wrthTne-BP.
 71121 Spike
the aelected aample
   with Tria-BP
                                                                                   7.1.1.1.1 Add
                                                                                 equivalent amount
                                                                                   of anhydroue
                                                                                     Na,SO.. .
                                                                                                           i
                                                                                                        7.1 1.2.2 Add
                                                                                                        CH2CI2, •••)
                                                                                                          •nd
                                                                                                         thr«« t
7.1.1.1
eampla
1
7.1
.3 Pack
m pipat.
r
1.1.4
Extract aampla
firat with CH3OH
followed by
                                              7 1.1.2.3
                                             Allow organic
                                              • nd water
                                               l*y«r« to
                                                                                     7.1.1.1.6
                                                                                   Reduce volum*
                                                                                    fay K-0 or N 2
                                                                                    Slowdown.
                                                                                                          7.1.1.2.4
                                                                                                         Collect the
                                                                                                          • xtrect
                                                             7.4 »«t HPLC/TS/MS
                                                                 operating
                                                                condition*.
                        7.6 Instrument
                         Cehbretion
                         procedure.
    7 fl Perform
   HPLC/TS/MS
    enelyeie.
                                                                                                           7 7 U»e
                                                                                                        Method 8000 to
                                                                                                       calculate enalyta(e)
                                                                                                        concentration.
                                                    8321A  -  50
                                                                                                    Revision  1
                                                                                                 January 1995

-------
                                  METHOD 8325

                SOLVENT EXTRACTABLE NON-VOLATILE COMPOUNDS BY
    HIGH PERFORMANCE LIQUID CHROMATOGRAPHY/PARTICLE BEAM/MASS SPECTROMETRY
                                  (HPLC/PB/MS)
1.0   SCOPE  AND APPLICATION

      1.1  This  method   describes   the   use   of  high   performance  liquid
chromatography  (HPLC), coupled with particle beam (PB) mass spectrometry  (MS),
for the determination of benzidines  and nitrogen-containing  pesticides  in water
and wastewater.  The following compounds can be determined  by this method:
            Compound                                    CAS No.'
            Benzidine                                    92-87-5
            Benzoylprop ethyl                         33878-50-1
            Carbaryl                                     63-25-2
            o-Chlorophenyl thiourea                    5344-82-1
            3,3'-Dichlorobenzidine                       91-94-1
            3,3'-Dimethoxybenzidine                     119-90-4
            3,3'-Dimethylbenzidine                      612-82-8
            Diuron                                      330-54-1
            Linuron (Lorox)                             330-55-2
            Monuron                                     150-68-5
            Rotenone                                     83-79-4
            Siduron                                    1982-49-6
a Chemical Abstract Services Registry Number

      1.2   The method also may be appropriate for the  analysis of benzidines and
nitrogen-containing pesticides  in non-aqueous  matrices.    The method  may  be
applicable to other compounds that can be extracted from a  sample with methylene
chloride and are amenable to separation on a reverse phase  liquid chromatography
column and transferable to the mass spectrometer with  a  particle beam interface.

      1.3   Preliminary  investigation indicates that the  following compounds also
may  be  amenable to this  method:   Aldicarb sulfone,  Carbofuran,  Methiocarb,
Methomyl (Lannate), Mexacarbate  (Zectran), and N-(l-Naphthyl)thiourea.  Ethylene
thiourea  and  o-Chlorophenyl   thiourea   have  been   successfully  analyzed  by
HPLC/PB/MS, but have not been successfully extracted  from a water matrix.

      1.4   Tables 4-6 present method detection  limits  (MDLs) for  the target
compounds,  ranging  from 2 to  25 M9/L.   The MDLs  are compound- and  matrix-
dependent.

      1.5   This  method  is restricted to  use by, or  under  the  supervision of,
analysts experienced in the  use of HPLC and skilled in  the interpretation  of

                                   8325 - 1                         Revision 0
                                                                  January 1995

-------
particle beam mass spectrometry.  Each analyst must demonstrate the ability to
generate acceptable results with this method.


2.0   SUMMARY  OF  METHOD

      2.1   The target compounds for this method must be extracted from the sample
matrix prior to analysis.

           2.1.1    Benzidines  and nitrogen-containing pesticides  are extracted
      from  aqueous matrices at  a neutral  pH with methylene chloride,  using a
      separatory  funnel   (Method 3510),  a  continuous  liquid-liquid  extractor
      (Method  3520), or  other  suitable technique.

           2.1.2    Solid samples are  extracted using Methods 3540 (Soxhlet), 3541
      (Automated  Soxhlet),  3550  (Ultrasonic  extraction),  or other  suitable
      technique.

      2.2   An  aliquot  of  the  sample  extract  is  introduced into  the  HPLC
instrument and a gradient elution program is used to chromatographically separate
the target analytes, using reverse-phase liquid chromatography.

      2.3   Once separated, the analytes are transferred to  the mass spectrometer
via a particle  beam HPLC/MS  interface.    Quantitation is performed  using an
external standard approach.

      2.4   An  optional internal  standard quantitation  procedure is included for
samples which  contain coeluting compounds or where  matrix interferences preclude
the use of the external  standard procedure.

      2.5   The use of  ultraviolet/visible  (UV/VIS) detection is an appropriate
option for the analysis of routine samples, whose general  composition has been
previously determined.


3.0   INTERFERENCES

      3.1   Refer   to  Methods   3500   and   8000  for   general   discussions  of
interferences  with  the  sample  extraction  and  chromatographic  separation
procedures.

      3.2  Although this  method relies on mass spectrometric detection, which can
distinguish between chromatographically  co-eluting compounds  on  the  basis of
their masses,  co-elution of two or more compounds will adversely affect method
performance.   When two  compounds coelute,  the  transport efficiency  of both
compounds through the particle  beam  interface  generally  improves,  and the ion
abundances observed in  the  mass spectrometer increase.   The  degree  of signal
enhancement by coelution is compound-dependent.

           3.2.1   This  coelution effect  invalidates the calibration curve and,
      if  not  recognized,  will  result in  incorrect  quantitative  measurements.
      Procedures  are given in this method  to check  for  co-eluting compounds, and
      must  be  followed to preclude inaccurate measurements.


                                   8325 - 2                         Revision 0
                                                                  January 1995

-------
           3.2.2   An optional  internal  standard calibration procedure has been
      included for use in instances of severe co-elution or matrix interferences.

      3.3   A major source of potential  contamination is HPLC columns which may
contain  silicon  compounds and   other  contaminants  that   could  prevent  the
determination of method analytes. Generally,  contaminants  will be leached from
the columns into mobile phase  and produce a variable background.  Figure  1 shows
unacceptable background contamination from a column with stationary phase bleed.

      3.4    Contamination  may  occur  when  a  sample  containing  low   analyte
concentrations is analyzed immediately after a sample containing relatively high
analyte concentrations.   After  analysis  of a sample  containing  high  analyte
concentrations, one or  more method  blanks  should  be analyzed.   Normally, with
HPLC, this is not a problem unless the  sample  concentrations are at the  percent
level.
4.0  APPARATUS  AND  MATERIALS

     4.1   High  performance  liquid chromatograph (HPLC)  - An analytical system
with programmable solvent delivery system and all necessary accessories including
5  nl injection  loop,  analytical  columns,  purging gases,  etc.    The  solvent
delivery system must be capable,  at  a  minimum, of handling  a  binary solvent
system, and must  be  able to  accurately  deliver flow rates between 0.20 - 0.40
mL/min.  Pulse dampening is  recommended,  but not required.  The chromatographic
system  must  be  able to  be  interfaced  with   a mass spectrometer  (MS).   An
autoinjector is recommended and should be capable of accurately delivering 1 -
10 juL injections  without affecting the chromatography.

           4.1.1  HPLC  Columns - An analytical column  is  needed,  and  a guard
     column  is  highly recommended.

                  4.1.1.1   Analytical   Column  -   Reverse  phase   column,   C18
           chemically bonded  to 4-10  JLUH silica particles, 150 -  200 mm x 2 mm,
           (Waters C-18 Novapak or equivalent).  Residual  acidic sites should be
           blocked (endcapped) with methyl or other non-polar groups  and the
           stationary phase  must  be  bonded to  the  solid  support  to  minimize
           column  bleed.  Select  a column that exhibits  minimal  bleeding.  New
           columns must  be  conditioned overnight before  use  by pumping a 75 -
           100%  v/v  acetonitrile:water solution  through the column at a rate of
           about 0.05 mL/min.   Other packings  and column  sizes may be  used if
           appropriate performance can be achieved.

                  4.1.1.2   Guard  Column - Packing  similar  to  that  used  in
           analytical column.

           4.1.2  HPLC/MS interface  - The particle beam HPLC/MS interface must
     reduce  the ion source  pressure  to  a level compatible with  the generation
     of classical electron ionization (El) mass spectra, i.e., about 1  x 10"4 -
     1  x  10"6 Torr,  while delivering sufficient quantities  of analytes to the
     conventional  El  source  to  meet  sensitivity,   accuracy,  and  precision
     requirements.   The concentrations  of background components  with masses
     greater  than 62 Daltons should be  reduced to  levels  that do  not  produce


                                   8325  - 3                         Revision 0
                                                                  January 1995

-------
      ions  greater  than  a  relative  abundance  of 10% in the mass spectra of the
      analytes.

      4.2   Mass  spectrometer  system -  The mass spectrometer must be capable of
electron ionization  at  a  nominal  electron energy of  70  eV.   The spectrometer
should  be  capable  of scanning  from  45  to  500  amu  in  1.5  seconds  or  less
(including scan overhead).  The spectrometer should produce a mass spectrum that
meets the criteria in Table 1 when  500 ng or  less of DFTPPO are introduced into
the HPLC.

      4.3   Data  system  -  A  computer  system  must  be  interfaced  to the  mass
spectrometer, and must be capable of the continuous acquisition and storage on
machine-readable media of all mass spectra obtained throughout the duration of
the chromatographic program.   The computer software must be capable of searching
any HPLC/MS data file for ions of a specified mass and plotting such abundance
data versus time or scan number.

      4.4   Volumetric  flasks   - Class  A,  in  various sizes, for  preparation of
standards.

      4.5   Vials - 10-mL amber glass vials with Teflon®-lined screw caps or crimp
tops.

      4.6   Analytical  balance -  capable  of weighing 0.0001  g.

      4.7   Extract  filtration apparatus

           4.7.1    Syringe  -  10-mL, with Luer-Lok  fitting.

           4.7.2    Syringe  filter assembly, disposable - 0.45 /^m pore size PTFE
      filter  in filter  assembly with Luer-Lok fitting  (Gelman Acrodisc,  or
      equivalent).


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests.  Unless otherwise
indicated,  it is intended  that all  reagents shall conform to the specifications
of the Committee on Analytical Reagents  of the American Chemical Society, where
such  specifications  are available.   Other grades may  be  used,  provided it is
first ascertained that the reagent  is  of sufficiently  high  purity to permit its
use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent  water  -  All  references to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Solvents  -  All  solvents  must  be HPLC-grade  or  equivalent.

           5.3.1    Acetonitrile,  CH3CN

           5.3.2    Methanol,  CH3OH

           5.3.3    Ammonium acetate, NH4OOCCH3, (0.01M in water).


                                   8325  - 4                         Revision 0
                                                                  January 1995

-------
      5.4   Mobile phase  -  Two mobile  phase  solutions  are  needed,  and  are
designated Solvent A and Solvent B.  Degas both solvents in an ultrasonic bath
under reduced pressure and maintain by purging with a low flow of helium.

           5.4.1    Solvent A   is  a water:acetonitrile  solution  (75/25,  v/v)
      containing  ammonium acetate  at  a  concentration of 0.01M.

           5.4.2    Solvent B is 100 % acetonitrile.

      5.5   Stock  standard solutions - Stock solutions may be prepared from pure
standard materials or purchased as certified solutions.  Commercially-prepared
stock standards  may be used  at any concentration  if  they are certified by the
manufacturer.

           5.5.1    Prepare stock standard solutions by accurately weighing 0.0100
      g  of pure material   in  a  volumetric flask.   Dilute to  known  volume in a
      volumetric  flask.   If compound purity is certified at 96% or greater, the
      weight may be used without correction to calculate the concentration of the
      stock standard.  Commercially-prepared  stock standards may be used at any
      concentration  if   they   are  certified  by   the  manufacturer   or  by  an
      independent source.

                   5.5.1.1  Dissolve    benzidines    and    nitrogen-containing
           pesticides in  methanol,  acetonitrile, or organic-free reagent water.

                   5.5.1.2  Certain analytes, such as 3,3'-dimethoxybenzidine,
           may require dilution in 50% (v/v) acetonitrile:water or methanol :water
           solution.

                   5.5.1.3  Benzidines may be used for calibration purposes in
           the free base or acid chlorides forms.  However, the concentration of
           the standard should  be calculated  as the free  base.

           5.5.2    Transfer the stock standard solutions into  amber bottles with
      Teflon®-!ined screw-caps  or crimp tops.  Store at -10°C or  less and protect
      from  light.  Stock  standard  solutions  should be checked  frequently for
      signs of  degradation or  evaporation, especially just prior  to  preparing
      calibration  standards from them.

      5.6   Surrogate  spiking   solution  -   The   recommended   surrogates  are
benzidine-D8, caffeine-  N2, 3,3'-dichlorobenzidine-D6, andbis-(perfluorophenyl)-
phenylphosphine  oxide.   Prepare  a solution  of  the surrogates in  methanol  or
acetonitrile  at  a  concentration of 5  mg/mL of each.   Other  surrogates may  be
included in this solution as needed.    (A 10-^L aliquot of  this solution  added
to 1 L of  water gives a concentration of  50 jiig/L of each surrogate).  Store the
surrogate  spiking solution in  an amber vial  in a  freezer  at  -10°C  or  less.

      5.7   MS performance check solution - Prepare a 100 ng//LtL  solution of DFTPPO
in acetonitrile.   Store this  solution  in  an amber vial in a freezer at -10°C or
less.
                                   8325 - 5                         Revision 0
                                                                  January 1995

-------
      5.8   Calibration  solutions

      This  method describes  two  types of  calibration  procedures that  may  be
applied to the target  compounds:   external  standard  calibration,  and internal
standard calibration.  Each procedure requires separate calibration standards.
In addition, the performance characteristics of the HPLC/PB/MS system indicate
that  it may  be  necessary to employ a second  order regression for calibration
purposes,  unless a very narrow calibration range is chosen.  See Method 8000 for
additional  information on non-linear calibration techniques.

      As described  in Method 8000, the analyst has two options for performing a
non-linear calibration.  The analyst may prepare five calibration standards and
perform triplicate  analyses  of each  standard,  or the  analyst may  prepare  10
calibration standards and perform a single analysis of each standard.

      Whichever approach  is used (a narrow linear range, the triplicate 5-point
approach,  or the single 10-point approach), the lowest  standard should be at a
concentration near, but above, the MDL of the target compound,  and the remaining
standards  should span the working range of the instrument.

           5.8.1    For  external   standard  calibration,   prepare  calibration
      standards for all target compounds  and surrogates  in acetonitrile.  DFTPPO
      may be  added  to  one or more  calibration solutions to  verify MS tune (see
      Sec.  7.3).  Store these solutions in amber vials at -10°C or less.   Check
      these solutions at  least quarterly for  signs of deterioration.

           5.8.2    Internal  standard  calibration requires the  use  of suitable
      internal  standards   (see  Method  8000).   Ideally,  stable,  isotopically-
      labeled,  analogs  of the target compounds should  be used.   These labeled
      compounds are included in the calibration standards  and must also be added
      to  each  sample  extract  immediately  prior  to  analysis.   Prepare  the
      calibration  standards  in  a  fashion  similar to that  for  external standard
      calibration,  but  include each internal standard  in each of the calibration
      standards.

           The concentration of the internal standards should be 50 - 100 times
      the lowest concentration of the unlabeled target compounds.  In addition,
      the  concentration  of  the  internal   standards  does  not  vary with  the
      concentrations of the target compounds,  but is held  constant.  Store these
      solutions in  amber vials at -10°C or  less.  Check these solutions at least
      quarterly for signs  of deterioration.

      5.9   Internal  standard  spiking solution  -  This  solution  is required when
internal  standard quantitation is used.   Prepare a solution containing each  of
the internal standards that will  be used  for quantitation  of  target compounds
(see Sec.  5.8.2) in methanol.  The concentration of this solution must be such
that a 1-juL  volume  of  the spiking  solution  added  to  a  1-mL  final  extract will
result in  a concentration  of each  internal  standard  that   is  equal  to  the
concentration of  the  internal  standard in  the calibration standards in Sec.
5.8.2.  Store  this solution in  an  amber vial at  -10°C  or less.   Check this
solution  at  least  quarterly for signs of  deterioration.  This solution  is not
necessary  if only external standard calibration will  be used.
                                   8325 - 6                         Revision 0
                                                                  January 1995

-------
      5.10  Sodium  chloride,  NaCl  -  granular,  used during  sample  extraction.

6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See the introductory material  to this Chapter, Organic Analytes, Sec.
4.1.

      6.2   Samples should be extracted within 7 days  and analyzed within 30 days
of extraction.   Extracts should be stored in amber vials at -10°C or less.


7.0   PROCEDURE

      7.1   Samples should be extracted by either Method 3510 (separatory funnel)
or Method 3520  (continuous  extractor)  or other appropriate technique.  Prior to
extraction, add a 10-/uL aliquot of the  surrogate spiking solution and 100 g of
sodium chloride to the  sample,  and  adjust the pH  of  the sample to 7.0.  Samples
of  other  matrices should  be  extracted  by  an appropriate sample  preparation
technique.  The concentration  of surrogates  in the sample should be 20-50 times
the method detection limit.  Concentrate the extract to 1 mL,  and exchange the
solvent to methanol,  following the procedures in  the extraction method.

      7.2   Establish   chromatographic,   particle   beam   interface,   and   mass
spectrometer conditions, using the following conditions as guidance.
      Mobile phase purge:
      Mobile phase flow rate:
      Gradient elution:
      Desolvation chamber temperature:
      Ion source temperature:
      Electron energy:
      Scan range:
Helium at 30 mL/min, continuous
0.25 - 0.3 mL/min through the column
Hold for  1  min at  25% acetonitrile
(Solvent A),  then program linearly to
about 70% acetonitrile (60% Solvent
B) in 29 min.  Start data acquisition
immediately.
45 - 80°C
250 - 290°C
70 eV
62 to 465 amu, at <1.5 sec/scan
NOTE: Post-column addition  is  an option  that  improves system  precision  and,
      thereby, may improve sensitivity.   Post-column  flow  rates depend on the
      requirements of  the  interface and may range  from 0.1 to  0.7  mL/min  of
      acetonitrile.   Maintain a minimum of 30% acetonitrile in the interface.

     Analyte-specific  chromatographic  conditions are  also shown in  Table  2.
(The particle beam interface conditions will  depend  on the  type of nebulizer).

          7.2.1   The  analyst  should  follow  the  manufacturer's  recommended
     conditions  for  their  interface's  optimum performance.   The interface  is
     usually  optimized  during  initial  installation  by flow  injection  with
     caffeine   or   benzidine,   and   should   utilize  a   mobile  phase   of
     acetonitrile/water  (50/50,  v/v).   Major  maintenance  may require  re-
     optimization.
                                   8325 - 7
                          Revision 0
                        January 1995

-------
     7.2.2   Fine tune the interface by making a series of injections into
the  HPLC  column  of  a  medium concentration  calibration  standard  and
adjusting the operating conditions (Sec.  7.2) until optimum sensitivity and
precision are obtained for the maximum number of target compounds.

7.3  Initial calibration

     7.3.1   Once the operating conditions have been established, calibrate
the MS  mass and abundance  scales  using DFTPPO  to  meet the  recommended
criteria in Table 1.

     7.3.2   Inject a medium concentration standard  containing DFTPPO,  or
separately  inject  into the  HPLC  a  5-/iL  aliquot  of  the 100 ng//uL  DFTPPO
solution and acquire a mass spectrum.  Use HPLC  conditions  that produce a
narrow  (at  least ten  scans  per peak) symmetrical peak.  If the  spectrum
does not meet the criteria (Table 1), the MS ion source must be  retuned and
adjusted to meet  all criteria  before proceeding with  calibration.   An
average  spectrum   across  the  HPLC   peak may  be used  to  evaluate  the
performance of the system.

     Manual  (not automated) ion source tuning  procedures specified by the
manufacturer should be employed during tuning.  Mass  calibration should be
accomplished while an  acetonitrile/water (50/50,  v/v) mixture  is  pumped
through the  HPLC column  and  the  optimized particle  beam interface.   For
optimum  long-term  stability  and  precision,  interface  and  ion  source
parameters  should be set near the  center of a  broad  signal  plateau rather
than at the peak  of a  sharp  maximum  (sharp maxima  exhibit  short-term
variations with particle  beam interfaces  and gradient elution conditions).

     7.3.3   System  performance  criteria  for the  medium concentration
standard -  Evaluate the stored HPLC/MS data with  the data system software
and verify  that the  HPLC/PB/MS  system  meets  the following  performance
criteria.

             7.3.3.1  HPLC   performance  -  3,3'-dimethylbenzidine   and
     3,3'-dimethoxybenzidine should be separated by  a  valley whose height
     is less than 25% of the average peak  height  of these  two compounds.
     If the valley  between them exceeds 25%, modify the gradient.  If this
     fails, the HPLC column requires maintenance.  See Sec. 7.4.6.

             7.3.3.2  Peak tailing  - Examine a total  ion chromatogram and
     examine the degree of peak tailing.  Severe tailing indicates a major
     problem and system  maintenance is  required to correct the  problem.
     See Sec. 7.4.6

             7.3.3.3  MS  sensitivity -  The signal-to-noise ratio for any
     compound's spectrum should be at least 3:1.

             7.3.3.4  Column  bleed  -  Figure  1  shows  an unacceptable
     chromatogram with column  bleed.  Figure 2  shows an  acceptable  ion
     chromatogram.    Figure  3  is  the mass  spectrum of dimethyloctadecyl-
     silanol, a common  stationary  phase bleed product.  If unacceptable
                              8325  -  8                         Revision  0
                                                             January 1995

-------
     column bleed is present,  the  column must be changed or conditioned to
     produce an acceptable background.

             7.3.3.5   Coeluting compounds - Compounds which coelute cannot
     be measured accurately because of carrier effects in the particle beam
     interface.  Peaks must be examined carefully for coeluting substances
     and  if  coeluting compounds  are present  at  greater  than  10%  of  the
     concentration  of the  target compound,  either conditions  must  be
     adjusted to resolve the components,  or internal  standard calibration
     must be used.

     7.3.4   Once optimized,  the same instrument operating conditions must
be used  for  the analysis of all  calibration  standards,  samples,  blanks,
etc.

     7.3.5   Once  all the performance criteria  are met,  inject a  5-juL
aliquot of each of the other calibration  standards using  the same HPLC/MS
conditions.

             7.3.5.1   The  general method of calibration is a second order
     regression  of integrated ion  abundances of  the  quantitation  ions
     (Table  3)  as  a  function  of  amount   injected.   For  second  order
     regression, a sufficient  number of calibration points must be obtained
     to accurately determine the  equation of the  curve.   (See Method 8000
     for the requirements for the number of  standards that must be employed
     for a non-linear calibration).   Non-linear calibration  models can be
     applied to either  the  external  standard or  the  internal  standard
     calibration approaches described here.

             7.3.5.2   For  some  analytes  the  instrument  response may  be
     linear over  a narrow concentration  range.   In  these  instances,  an
     average calibration factor  (external  standard)  or average  response
     factor (internal  standard) may  be employed  for sample  quantitation
     (see Method 8000).

     7.3.6   If  a  linear calibration model  is used,  calculate  the  mean
calibration factor or response factor for each  analyte,  including  the
surrogates, as  described  in Method 8000.  Calculate the  standard deviation
(SD) and  the relative standard  deviation  (RSD) as well.   The RSD  of  an
analyte or surrogate must be less than or  equal  to 20%, if the linear model
is to be applied.  Otherwise,  proceed as  described in Method 8000.

7.4  Calibration verification

     Prior to sample analysis, verify the MS tune  and initial  calibration
at  the  beginning  of  each 8-hour  analysis  shift  using  the  following
procedure:

     7.4.1   Inject a  5-juL aliquot of  the  DFTPPO  solution or a mid-level
calibration standard  containing  500  ng  of  DFTPPO,  and  acquire a mass
spectrum that includes data for m/z 62-465.  If the spectrum  does  not meet
the criteria in  Table  1, the MS must be  retuned to meet the criteria before
proceeding with the continuing calibration  check.
                              8325  -  9                          Revision 0
                                                             January  1995

-------
           7.4.2    Inject  a  5-/xL aliquot of a medium concentration calibration
      solution  and analyze  with  the same  conditions  used during  the  initial
      calibration.

           7.4.3    Demonstrate  acceptable performance for the criteria shown in
      Sec.  7.3.3.

           7.4.4    Using the initial  calibration (either linear or non-linear,
      external  standard or internal  standard),  calculate the  concentrations in
      the medium  concentration  calibration  solution  and compare the results to
      the known values in the calibration solution.  If calculated concentrations
      deviate  by  more than  20% from known  values,  adjust the  instrument  and
      inject the  standard  again.  If the calibration  cannot be verified with the
      second injection, then a  new initial  calibration  must  be performed after
      taking corrective actions  such  as  those described  in Sec. 7.9.

      7.5   Sample Analysis

           7.5.1    The column should be conditioned overnight before each use by
      pumping  a acetonitrile:water (70% v/v) solution through  it  at  a  rate of
      about 0.05  mL/min.

           7.5.2    Filter  the extract through  a 0.45  jum filter.   If internal
      standard  calibration  is  employed, add  10 juL of the internal  standard
      spiking  solution to  the 1-mL final extract immediately before injection.

           7.5.3    Analyze a 5-/iL aliquot of the  extract,  using  the  operating
      conditions  established in  Sees. 7.2 and 7.3.

      7.6   Qualitative identification

      The qualitative  identification  of compounds  determined by this  method is
based on retention time and on comparison of  the sample  mass  spectrum,  after
background correction, with characteristic  ions in  a  reference mass  spectrum.
The  reference  mass spectrum  must be  generated by the laboratory  using  the
conditions of  this method.   The  characteristic ions  from the  reference  mass
spectrum are defined as the three  ions  of  greatest  relative  intensity,  or any
ions  over  30%  relative  intensity, if  less  than three  such ions  occur  in  the
reference spectrum.  Compounds  are  identified  when  the following criteria are
met.

           7.6.1    The intensities of the characteristic ions of a compound must
      maximize  in the same scan or within one scan of each other.   Selection of
      a peak by a data system target  compound search  routine where the search is
      based  on the presence  of a target chromatographic peak  containing  ions
      specific  for the target compound at a compound-specific retention time will
      be accepted as meeting this  criterion.

           7.6.2    The retention  time  of the sample component is within ± 10%
      of the retention time  of  the standard.

           7.6.3    The  relative intensities  of the  characteristic ions  agree
      within  20%  of the  relative intensities  of  these ions  in  the  reference


                                  8325  -  10                         Revision 0
                                                                  January 1995

-------
spectrum.  (Example:  For an  ion with an abundance of 50% in the reference
spectrum,  the corresponding  abundance  in  a  sample  spectrum can  range
between 30% and 70%.)

     7.6.4    Structural  isomers  that produce  very  similar mass  spectra
should  be identified  as individual  isomers  if  they have  sufficiently
different HPLC retention  times.  Sufficient GC  resolution  is  achieved  if
the height of the valley between two isomer peaks is less than 25% of the
sum of the two peak  heights.  Otherwise, structural isomers are identified
as isomeric pairs.

     7.6.5    Identification  is  hampered when  sample  components  are  not
resolved  chromatographically and  produce  mass  spectra  containing  ions
contributed by more  than one  analyte.  When HPLC peaks  obviously represent
more than one sample component  (i.e., a broadened peak with shoulder(s)  or
a valley  between  two or  more  maxima),  appropriate  selection  of  analyte
spectra and background spectra is important.

     7.6.6    Examination  of extracted ion current profiles of appropriate
ions can aid  in the selection of spectra, and in qualitative identification
of compounds.  When  analytes coelute (i.e.,  only one chromatographic peak
is apparent),  the identification criteria may  be met,  but each  analyte
spectrum  will  contain  extraneous   ions  contributed  by  the  coeluting
compound.

7.7  Quantitative Analysis

     7.7.1    Complete chromatographic resolution is necessary for accurate
and  precise   measurements  of analyte  concentrations.   Compounds  which
coelute cannot be measured  accurately  because of carrier effects  in  the
particle beam interface.  Peaks must be examined  carefully  for coeluting
substances and if coeluting compounds are present at  greater  than 10%  of
the  concentration of  the target  compound,  either  conditions  must  be
adjusted to resolve  the components, or the results for the target compound
must be flagged as potentially positively biased.

     7.7.2    Calculate the concentration of each analyte, using either the
external standard or internal standard calibration.    See Method  8000  for
the specific equations to be employed for  either the non-linear or linear
calibration models.

     7.7.3    If the response for any quantitation ion  exceeds  the initial
calibration range of  the HPLC/PB/MS system,  the  sample  extract  must  be
diluted and reanalyzed.   When internal  standard calibration  is  employed,
additional internal  standard must  be  added  to the  diluted  extract  to
maintain the same concentration as  in the calibration  standards.

7.8  HPLC-UV/VIS Detection (optional)

     7.8.1    Prepare calibration solutions as outlined in Sec. 5.8.

     7.8.2    Inject 5 juL of each calibration solution onto the HPLC,  using
the  chromatographic  conditions outlined   in  Sees.  7.2.1   and   7.2.2.
                             8325 - 11                        Revision 0
                                                            January  1995

-------
      Integrate  the area  under  the full  chromatographic  peak at  the  optimum
      wavelength  (or at 230 nm if that option is not available) for each target
      compound at  each concentration.

           7.8.3    The retention time of the chromatographic peak is an important
      criterion  for  analyte  identification.    Therefore,  the ratio  of  the
      retention time of the sample analyte to  the standard  analyte should be 1.0
      ± 0.1.

           7.8.4    Calculate calibration factors or response factors as described
      in  Method  8000,  for  either  external   standard  or  internal  standard
      calibration,  and evaluate the calibration  linearity as described in Method
      8000.

           7.8.5    Verify the  calibration  at  the  beginning  of  each  8-hour
      analytical  shift, as described above.

           7.8.6  Once the calibration has been verified,  inject a  5-/A aliquot
      of the sample extract,  start the HPLC gradient elution,  load and inject the
      sample  aliquot,  and begin  data  acquisition.   Refer to Method 8000  for
      guidance on  calculation of concentration.

      7.9   Corrective Actions

      When  the initial  calibration cannot  be verified,  one  or  more of  the
following corrective actions may be necessary.

           7.9.1    Major maintenance such as cleaning an ion source, cleaning the
      entrance  lens,  quadrapole  rods,  etc.,  will   require  a  new  initial
      calibration.

           7.9.2    Check  and  adjust  HPLC and/or MS operating conditions;  check
      the MS  resolution, and calibrate the mass scale.

           7.9.3    Replace the mobile phases with fresh solvents. Verify that the
      flow  rate from the HPLC pump is constant.

           7.9.4    Flush  the HPLC column with acetonitrile.

           7.9.5    Replace the HPLC column.  This action will cause a change in
      retention times.

           7.9.6    Prepare fresh  calibration  solutions,  and repeat  the initial
      calibration  step.

           7.9.7    Replace any components that  leak.

           7.9.8    Replace the  MS  electron  multiplier,   or  any  other  faulty
      components.

           7.9.9    Clean  the interface to eliminate  plugged  components and/or
      replace skimmers according to the  manufacturer's instructions.
                                  8325  -  12                         Revision 0
                                                                  January 1995

-------
           7.9.10  If peak  areas are  determined  by the  instrument software,
      verify values by manual  integration.

           7.9.11  Increasing  ion source temperature can reduce peak tailing, but
      excessive  ion source temperature can affect the quality of the  spectra for
      some  compounds.

           7.9.12  Air leaks into the  interface may effect  the quality of the
      spectra (e.g.  DFTPPO),  especially when  the  ion  source is  operated  at
      temperatures  in  excess of  280°C.


8.0   QUALITY CONTROL

      8.1  Refer to Chapter One and Method 8000 for specific quality control (QC)
procedures.   Quailty  control  procedures  to  ensure the  proper operation of the
various  sample  preparation techniques  can  be  found in Method  3500.     Each
laboratory  should  maintain  a formal quality assurance program.  The laboratory
should also maintain records to  document the quality of the data generated.

      8.2    Quality control  procedures necessary  to evaluate  the  HPLC system
operation are found in Method 8000,  Sec. 7.0 and includes evaluation of retention
time windows, calibration verification and chromatographic analysis  of samples.
Necessary instrument QC is  found in the following sections.

           8.2.1    The HPLC/PB/MS  system  should be  tuned  to  meet  the DFTPPO
      criteria in Sees.  7.3.1  and 7.4.1.

           8.2.2    There should  be  an initial  calibration  of the  HPLC/PB/MS
      system as  described  in Sec. 7.3.

           8.2.3    The HPLC/PB/MS  system  should  meet  the  system  performance
      criteria in Sec. 7.3.3,  each  8 hours.

      8.3     Initial   Demonstration  of  Proficiency  -   Each  laboratory  must
demonstrate initial proficiency with each sample preparation and  determinative
method combination  it utilizes,  by generating  data of  acceptable accuracy and
precision for target analytes in a clean matrix.  The laboratory must  also repeat
the following operations whenever new staff are trained or significant changes
in instrumentation are made.  See Method 8000,  Sec. 8.0 for information on how
to accomplish this demonstration.

      8.4  Sample Quality Control for Preparation and Analysis - The laboratory
must  also have  procedures  for documenting  the effect of the  matrix on method
performance  (precision,  accuracy,  and detection  limit).   At  a  minimum,  this
includes the analysis of QC  samples including a method blank, a matrix spike,  a
duplicate,  and a  laboratory control  sample (LCS) in each  analytical batch and the
addition of surrogates to each field  sample  and QC sample.

         ' 8.4.1    Documenting  the effect  of  the matrix  should include  the
      analysis of at least one  matrix  spike and one duplicate unspiked sample  or
      one matrix  spike/matrix spike duplicate pair.  The decision  on whether  to
      prepare  and  analyze  duplicate   samples  or  a  matrix   spike/matrix  spike
      duplicate must be based on  a knowledge  of the samples in the  sample batch.

                                  8325 - 13                         Revision  0
                                                                  January 1995

-------
      If samples are expected to contain target analytes, then laboratories may
      use one matrix spike and a duplicate  analysis of an unspiked field sample.
      If samples are not expected to contain target analytes, laboratories should
      use a matrix  spike and matrix spike  duplicate pair.

          8.4.2  A Laboratory Control  Sample (LCS) should be included with each
      analytical  batch.   The LCS consists  of an aliquot of  a  clean  (control)
      matrix similar to the sample matrix and of  the same weight or volume.  The
      LCS  is  spiked with the  same  analytes at the same concentrations  as the
      matrix spike.   When the results of  the  matrix  spike  analysis  indicate a
      potential problem due to the sample matrix itself,  the  LCS results are used
      to verify that the  laboratory can perform the analysis in a clean matrix.

          8.4.3   See Method  8000,  Sec. 8.0  for  the details on  carrying out
      sample quality control procedures  for  preparation and analysis.

      8.5  Surrogate recoveries  - The laboratory must evaluate surrogate recovery
data from individual samples versus the surrogate control limits developed by the
laboratory.   See Method 8000, Sec.  8.0 for information  on  evaluating surrogate
data and developing and updating surrogate limits.

      8.6    It  is  recommended  that  the  laboratory  adopt  additional  quality
assurance  practices for use with this method.  The specific practices that are
most productive depend  upon the needs of  the  laboratory and  the  nature  of the
samples.   Whenever possible, the  laboratory should  analyze standard reference
materials  and participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1  Single laboratory accuracy  and precision data for the benzidines and
nitrogen-containing pesticides are  presented in Tables 4-6.  Five to seven 1-L
aliquots of organic-free reagent water,  containing approximately five times the
MDL of each  analyte, were analyzed with this procedure (Reference  1).  The final
extract volume was 0.5 mL for these determinations.

          9.1.1  Method detection  limits  (MDLs) are presented in Tables 4-6.

          9.1.2    A  multi-laboratory  (12  laboratories)  validation  of  the
      determinative  step was done  for four of the analytes  (benzidine,  3,3'-
      dimethoxybenzidine,   3,3'-dimethylbenzidine,   3,3'-dichlorobenzidine).
      Table  7  provides  the  results  from  this  study   for single  laboratory
      precision, overall  laboratory precision, and overall  laboratory accuracy.
      The  two  concentration  levels shown  represent  the two  extremes  of the
      concentration range studied.
                                   8325  -  14                         Revision 0
                                                                  January 1995

-------
10.0 REFERENCES

1.   Bellar,  T.A.,  Behymer,  T.D.,  Ho,  J.S.,  Budde,  W.L.,  "Method  553:
     Determination of Benzidines and Nitrogen-Containing Pesticides  in Water by
     Liquid-Liquid Extraction or Liquid-Solid Extraction and  Reverse Phase High
     Performance  Liquid Chrotnatography/Particle Beam/Mass Spectrometry",  U.S.
     Environmental  Protection Agency,  EMSL-Cincinnati, Revision  1.1,  August
     1992.

2.   Bellar,  T.A., Behymer,  T.D.,  Budde,  W.L.,  "Investigation of Enhanced Ion
     Abundances from a Carrier Process in High-Performance Liquid Chromatography
     Particle Beam  Mass Spectrometry", J.  Am.  Soc. Mass Spectrom.,  1990,  1,
     92-98.

3.   Behymer,     T.D.,    Bellar,    T.A.,    and     Budde,    W.L.,    "Liquid
     Chromatography/Particle  Beam/Mass  Spectrometry  of Polar  Compounds  of
     Environmental Interest", Anal.  Chem.,  1990, 62, 1686-1690.

4.   Ho, J.S., Behymer,  T.D.,  Budde,  W.L., and Bellar, T.A., "Mass Transport and
     Calibration  in  Liquid  Chromatography/  Particle Beam/  Mass Spectrometry",
     J. Am. Soc.  Mass Spectrom., 1992,  3, 662-671.
                                  8325  -  15                         Revision 0
                                                                  January 1995

-------
                                    TABLE  1

        ION ABUNDANCE CRITERIA FOR BIS(PERFLUOROPHENYL)PHENYLPHOSPHINE
                 (DECAFLUOROTRIPHENYLPHOSPHINE OXIDE,  DFTPPO)
m/z
Relative Abundance
Purpose of Specification1
77
168
169
271
365
438
458
459

Present, major ion
Present, major ion
4 - 10% of 168
Present, major ion
5 - 10% of base peak
Present
Present
15 - 24% of mass 458

Low mass sensitivity
Mid-mass sensitivity
Mid-mass resolution and isotope ratio
Base peak
Baseline threshold check
Important high mass fragment
Molecular ion
High mass resolution and isotope
ratio
      The primary use of all the  ions  is  to  check  the  mass  calibration of the
      mass spectrometer.  The second use of these ions  are the mass resolution
      checks,  including  the natural  isotope abundance  ratios.    The  correct
      setting  of  the  baseline  threshold  is  indicated  by the presence  of low
      intensity ions,  and  is the  third  use  of  this test.    Finally,  the ion
      abundance  ranges  may provide   some   standardization  to  fragmentation
      patterns of the target compounds.
                                   TABLE 2

                  RECOMMENDED  HPLC  CHROMATOGRAPHIC CONDITIONS
               FOR BENZIDINES  AND NITROGEN-CONTAINING  PESTICIDES
Initial Mobile       Initial
Phase (v/v %)      Time (min)
                          Gradient
                            Time
            Final  Mobile
           Phase  (v/v %)
75/25
(water1/CH3CN)
                             29
               30/70
           (waterVCH3CN)
    1   Water contains 0.01M ammonium acetate.
                                   8325  -  16
                                                        Revision 0
                                                      January 1995

-------
                                    TABLE  3

        RETENTION TIME DATA AND QUANTITATION IONS FOR TARGET COMPOUNDS
Compound
Benzidine
Benzoylprop ethyl
Caffeine
Carbaryl
0-Chlorophenyl thiourea
3,3'-Dichlorobenzidine
3,3' -Dimethoxybenzi dine
3, 3' -Dimethyl benzi dine
Diuron
Ethylene thiourea
Linuron
Rotenone
Siduron
Retention
Time
System la
4.3
24.8
1.4
10.1
2.7
16.6
8.1
8.5
11.0
1.2
16.0
21.1
14.8
Retention
Time
System 2b
4.9
31.3
1.6
14.7
3.0
22.7
11.5
12.4
16.1
1.4
21.9
27.4
20.6
Quantitation
Ion
184
105
194
144
151
252
244
212
72
102
161
192
93
Surrogates:'
Benzidine-d8                 4.2
Caffeine-15N2                 1.3
3,3'-Dichlorobenzidine-d6   16.5
bi s(Perf1uorophenyl)-
  phenylphosphine oxide     22.0
 4.8
 1.6
22.6

28.9
192
196
258

271
      These  retention  times were  obtained on  a  Hewlett-Packard  1090  liquid
      chromatograph with a Waters C18 Novapak 15 cm x  2 mm column using gradient
      conditions given in Table 1.

      These  retention  times  were  obtained   on  a Waters   600  MS  liquid
      chromatograph with a Waters C18 Novapak 15 cm x  2 mm column using gradient
      conditions given in Sec.  7.2.

      These compounds cannot be used as  surrogates if their unlabeled analogs
      are present (see Sec. 3.2).
                                  8325 - 17
                      Revision 0
                    January 1995

-------
                                   TABLE  4

 ACCURACY AND PRECISION DATA FROM SIX DETERMINATIONS OF THE TARGET COMPOUNDS
         IN ORGANIC-FREE REAGENT WATER USING LIQUID-LIQUID EXTRACTION



Compound
Benzidine
Benzoylprop ethyl
Caffeine
Carbaryl
o-Chlorophenyl thiourea
3,3'-Dichlorobenzidine
3,3'-Dimethoxybenzidine
3,3'-Dimethylbenzidine
Diuron
Ethylene thiourea
Linuron
Monuron
Rotenone
Siduron

True
Cone.
(M9/L)
22.9
32.5
14.4
56.6
32.6
24.8
31.6
31.7
25.0
32.0
95.0
31.2
50.3
27.9
Mean
Observed
Cone.
(M9/L)
20.5
33.0
10.5
52.2
15.3
21.7
29.2
31.8
26.2
0.0
89.5
31.8
44.9
29.6

Std.
Dev.
(M9/L)
0.8
1.1
0.9
2.9
2.2
0.7
2.3
1.0
1.3
0.0
3.9
1.2
9.4
1.4


RSD

3.3
3.3
6.3
5.1
6.8
2.9
7.3
3.1
5.1
0.0
4.1
3.8
18.8
5.2
Mean
Accuracy
(% of
True)
89.6
101.6
72.6
92.3
47.0
89.6
92.3
100.4
104.8
0.0
94.2
101.9
89.3
106.3


MDL
(M9/L)
2.5
3.7
3.1
9.8
7.4*
2.4
7.7
3.3
4.4
*
13.1
4.0
31.6
4.7
* Not recovered
                                   8325  -  18
  Revision 0
January 1995

-------
                                    TABLE 5

 ACCURACY AND PRECISION DATA FROM SEVEN DETERMINATIONS OF THE TARGET COMPOUNDS
  IN ORGANIC-FREE REAGENT WATER USING SOLID-PHASE EXTRACTION (C18 CARTRIDGE)8



Compound
Benzidine
Benzoylprop ethyl
Caffeine
Carbaryl
o-Chlorophenyl thiourea
3,3'-Dichlorobenzidine
3,3'-Dimethoxybenzidine
3,3' -Di methyl benzi dine
Diuron
Ethylene thiourea
Linuron
Monuron
Rotenone
Siduron

True
Cone.
(M9/L)
22.9
32.5
14.4
56.6
32.6
5.0
31.6
31.7
25.0
32.0
95.0
31.2
50.3
27.9
Mean
Observed
Cone.
(M9/L)
12.2
29.3
6.4
53.9
0.0
4.4
25.5
31.4
24.4
0.0
88.9
30.5
45.0
24.8

Std.
Dev.
(M9/L)
1.7
2.0
1.4
1.8
0.0
0.4
1.8
1.0
1.4
0.0
4.8
2.9
2.4
2.0


RSD
(%)
13.7
6.9
21.4
3.3
0.0
10.0
7.1
3.1
5.6
0.0
5.4
9.6
5.4
7.9
Mean
Accuracy
(% of
True)
53.2
90.2
44.2
95.2
0.0
89.6
80.8
99.0
97.6
0.0
93.6
97.8
89.6
88.9


MDL
(M9/L)
5.3
6.3
4.4
5.7
*
1.4
5.7
3.0
4.4
*
15.1
9.1
7.5
6.3
3  Reagent  water contained 0.01  M ammonium acetate.

*  Not recovered.
                                   8325  -  19
  Revision 0
January 1995

-------
                                    TABLE  6

 ACCURACY  AND  PRECISION  DATA  FROM  SIX  DETERMINATIONS OF THE TARGET  COMPOUNDS
          IN ORGANIC-FREE REAGENT WATER USING SOLID-PHASE EXTRACTION
               (NEUTRAL POLYSTYRENE/DIVINYLBENZENE  POLYMER DISK)



Compound
Benzidine
Benzoylprop ethyl
Caffeine
Carbaryl
o-Chlorophenyl thiourea
3,3'-Dichlorobenzidine
3,3'-Dimethoxybenzidine
3,3'-Dimethylbenzidine
Diuron
Ethylene thiourea
Linuron
Monuron
Rotenone
Siduron

True
Cone.
(M9/L)
22.9
32.5
14.4
56.6
32.6
5.0
31.6
31.7
25.0
32.0
95.0
31.2
50.3
27.9
Mean
Observed
Cone.
(M9/L)
24.7
31.1
0.7
59.5
0.0
5.0
32.8
31.5
26.1
0.0
97.9
34.4
40.5
26.8

Std.
Dev.
(M9/L)
2.4
3.0
0.5
4.7
0.0
0.5
2.2
2.1
1.8
0.0
8.7
2.5
6.0
1.0


RSD
(%)
9.8
9.6
72.5
7.9
0.0
9.4
6.7
6.7
7.0
0.0
9.0
7.3
14.8
3.6
Mean
Accuracy
(% of
True)
108.0
95.8
5.2
105.1
0.0
101.7
103.8
99.4
104.5
0.0
103.0
110.4
80.5
96.1


MDL
(M9/L)
8.1
10.1
1.8
15.8
*
1.6
7.4
7.1
6.1
*
29.3
8.4
20.2
3.4
* Not recovered.
                                   8325  -  20
  Revision 0
January 1995

-------
                                    TABLE  7

  MEAN  RECOVERIES,  MULTI-LABORATORY  PRECISION AND  ESTIMATES OF SINGLE ANALYST
         PRECISION FOR THE MEASUREMENTS OF FOUR BENZIDINES BY LC/PB/MS
                           10 uq/L Test Cone.       100 uq/L  Test  Cone.

                                   RSD       RSD                RSD       RSD
                       Recovery  Multi-   Single  Recovery  Multi-    Single
Compound                  (%)       lab    Analyst    (%)       lab    Analyst


Benzidine                 96       10       5.6      97      10        9.1
3,3'-Dimethoxybenzidine  104       20      18        95      10        7.0
3,3'-Dimethylbenzidine    98       14      10        97       8.6      4.9
3,3'-Dichlorobenzidine    96       18       9.4      97       9.1      4.6
                                  8325 - 21                         Revision  0
                                                                  January  1995

-------
                           FIGURE 1
      AN UNACCEPTABLE  CHROMATOGRAM WITH COLUMN BLEED
  COLUMN FOLLOWING EXPOSURE TO  AMMONIUM  ACETATE
 80000-
 48000-
 40000-
 38000-
 30000-
 23000-
 coooo-
 18000-
 10000-
  sooo-
     0-
OIMETHYLOCTADECYL3ILANOL
(MAJOR IONS  - M/Z  76,  313)
                         1C     it     20     C4     C«     3C
 100
 90
-•0
 79
-40
40
-40
 30
•CO
 10
-0
                           FIGURE 2
  AN ACCEPTABLE CHROMATOGRAM FOLLOWING COLUMN FLUSHING
CIS COLUMN  MAINTAINED WITH  ACETONITRILE  FLUSHING
  24000-

  20000-

  16000-

  12000-

   4000-

   4000-

     0-
                                                100
                                                »0
                                               •40
                                                70
                                                «0
                                               -BO
                                               -40
                                                30
                                               -CO
                                                10
     "k     ic
                                       CO    24     29     3C
                           8325  -  22
                                                  Revision 0
                                                January  1995

-------
                           FIGURE 3
            MASS SPECTRUM OF DIMETHYLOCTADECYL-SILANOL,
              A COMMON STATIONARY PHASE BLEED PRODUCT
            MASS SPECTRUM  OF  C18  COLUMN BLEED
100X     75
INT
   I '
   50
                     CH.
HO-Si-Ci
     I
    CHo
                                     DIMETHYLOCTADECYLSILANOL
                                                     M-Methy

                                                        313
                                                 281
150
200
250
                                    390
                                M/Z
                           8325 - 23
                                       Revision 0
                                     January 1995

-------
                                       FIGURE  4

                  TOTAL  ION CHROMATOGRAM OF ANALYTES AND SURROGATES
                                (140-950 ng Injected)
1200000
        160  300   300  400   500  640
    O.3ML/MIN  O.O1 M AMMONIUM ACETATE
               1100 1200  1300 1400



POST COLUMN 0.1 M L / M I N  ACETONITRILE
                                       8325 - 24
                         Revision 0
                       January 1995

-------
                                 METHOD 8325
               SOLVENT EXTRACTABLE NON-VOLATILE COMPOUNDS  BY
  HIGH PERFORMANCE LIQUID CHROMATOGRAPHY/PARTICLE BEAM/MASS  SPECTROMETRY
                                 (HPLC/PB/MS)
     7.1  Choose
appropriate extraction
      technique.
7.6 Perform qualitative
   identification of
   compounds with
 appropriate reference
      spectrum.
     7.2 Establish
   chromatographic,
    particle beam
interference, and mass
    spectrometer
      conditions.
                                                     7.7 Calculate
                                                   concentrations of
                                                     each sample.
                                           Yes
     7.3 Calibrate
      equipment.
                              7.7.3 Dilute
                                sample.
      7.4 Verify
     calibration.
     7.5 Perform
   sample analysis.
       7.7.3
       Does
  response exceed
   the calibration
       range?
        7.8
         HPLC:
      UV/UIS
    detection be
    performed?
     (Optional)
                                                     7.8.1 - 7.8.4
                                                  Calibrate equipment
                                                      7.8.5 Verify
                                                      calibration.
                                   8325 - 25
                      Revision  0
                    January  1995

-------
                                  METHOD 8332

           NITROGLYCERINE BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
1.0 SCOPE AND APPLICATION

      1.1 Method 8332 is used to determine the concentration of nitroglycerine
(CAS No.  55-63-0)  in aqueous matrices  such  as waste water,  groundwater,  and
surface water.  This method may  also  be applicable  to other matrices.   Method
8332 also  provides a  qualitative  procedure  for  determining the  presence  or
absence of nitroglycerine (See Appendix A).

      1.2   This method is restricted to  use  by,  or under the supervision of,
analysts experienced  in the  use of HPLC  instrumentation and skilled  in  the
interpretation of HPLC  chromatograms.  Each analyst must demonstrate the ability
to generate acceptable results with this method.


2.0 SUMMARY OF METHOD

      2.1  Samples  are  injected  onto a reverse phase  high  performance liquid
chromatograph (HPLC) column  and the nitroglycerine concentration is quantitated
using a UV detector.


3.0 INTERFERENCES

      3.1   Solvents, reagents,  glassware and other sample processing hardware
may yield discrete artifacts  and/or elevated baselines, causing misinterpretation
of the chromatograms.  All of these materials must  be demonstrated to  be free
from interferences.
4.0 APPARATUS AND MATERIALS

      4.1  HPLC system

            4.1.1 HPLC (isocratic) - equipped with a  pump,  a  direct injection
      port or 20 nl loop injector and a 214 nm UV detector.

            4.1.2  Column  -  Waters  Radial -  Pak  CN,  10 nm particle  size (or
      equivalent).

            4.1.3  Integrator.

      4.2  Injection  syringe.


5.0 REAGENTS

      5.1 Reagent grade chemicals  shall be used in all tests.  Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications


                                  8332 -  1                          Revision  0
                                                                  January 1995

-------
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications  are  available.   Other grades may be  used,  provided it is
first ascertained that the reagent is  of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2  Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3  Acetonitrile, CH3CN.

      5.4    Nitroglycerine   -  "Nitrostat"  tablets   (nominal  concentration  =
0.4 mg/tablet),  available from  Warner  Lambert/Parke Davis.    Certificate  of
Analysis for  any current lot  is  available  from Warner Lambert,  Lititz,  Pa.
(717) 626-2011.

      5.5  HPLC mobile phase  eluant  -  60% Acetonitrile/40% Organic-free reagent
water.

      5.6  HPLC calibration  standards  -  Dissolve five nitroglycerine tablets in
the HPLC mobile phase, using a 500 mL volumetric flask.   Dilute to volume with
mobile phase.   The resulting standard will contain 4 mg/L nitroglycerine.  8 mg/L
and 12 mg/L nitroglycerine standard solutions can be prepared in the same manner.
Other concentrations, as needed,  can be  prepared by appropriate dilutions.  Use
the concentration listed on the "Nitrostat" Certificate of Analysis to calculate
the concentration of the standards to three significant figures.


6.0 SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this Chapter,  Organic Analytes,
Section 4.1.
7.0 PROCEDURE

      7.1  Chromatographic Conditions (Recommended):

      Flow rate:  1.0 mL/min.
      UV Detector:    214 nm.
      Injection:  20 /zL  (for 4 to 12 mg/L nitroglycerine).

      Chromatographic conditions should  be established  to give a retention time
for nitroglycerine of approximately 5.1  min.

      7.2    Initial  Calibration  -  Prepare  a  5-point  calibration curve  and
establish the  calibration range of  the  method  (nominally between  4  mg/L  and
12 mg/L).  Calculate a correlation coefficient,  slope,  and zero intercept from
the regression analysis of the data points  (peak area vs.  concentration).

      7.3  Make  20  ^L injections for each  sample or sample  dilution.   If the
response for the  sample exceeds that  of the initial  calibration range, the sample
must  be  diluted.   Calculate sample  concentrations using  the  slope  and  the
intercept figures from the regression analysis of the standards.


                                   8332  - 2                         Revision 0
                                                                  January 1995

-------
8.0  QUALITY CONTROL

      8.1  Refer to  Chapter One and Method 8000 for specific quality control  (QC)
procedures.  Each laboratory should maintain a formal  quality assurance program.
The laboratory should also maintain records to document the quality of the  data
generated.

      8.2   Quality control  procedures  necessary to evaluate  the HPLC system
operation are found in Method 8000, Sec.  7.0 and includes evaluation of retention
time windows, calibration verification and  chromatographic analysis of samples.

      8.3    Initial  Demonstration  of  Proficiency   -  Each   laboratory   must
demonstrate  initial proficiency with  each  sample preparation and determinative
method combination  it  utilizes,  by generating data of acceptable  accuracy and
precision for target analytes  in a clean matrix.  The laboratory must also repeat
the following operations whenever  new staff are trained or  significant changes
in instrumentation  are made.  See  Method 8000, Sec. 8.0 for information on how
to accomplish this  demonstration.

      8.4  Sample Quality Control  for Preparation and Analysis  - The laboratory
must also  have  procedures  for documenting the effect  of  the matrix on method
performance  (precision,  accuracy,  and  detection limit).    At  a  minimum,   this
includes the analysis of QC samples including  a method  blank, a matrix spike, a
duplicate, and a laboratory control sample (LCS) in each  analytical batch.

            8.4.1   Documenting the  effect of the  matrix  should  include the
      analysis of at least  one matrix  spike and one  duplicate unspiked sample or
      one matrix spike/matrix spike duplicate  pair.   The decision  on whether to
      prepare  and   analyze  duplicate  samples  or  a matrix  spike/matrix  spike
      duplicate must be based on a  knowledge of the  samples in the  sample batch.
      If samples are expected to  contain target analytes,  then  laboratories may
      use one matrix spike  and a duplicate analysis  of an unspiked  field sample.
      If  samples are  not  expected to  contain  target  analytes,   laboratories
      should use a  matrix spike and matrix spike duplicate pair.

            8.4.2   A  Laboratory  Control Sample (LCS)  should  be  included with
      each  analytical  batch.   The  LCS consists  of  an   aliquot  of a  clean
      (control) matrix similar to  the sample  matrix  and  of the same weight or
      volume.    The  LCS  is  spiked   with the  same  analytes  at  the   same
      concentrations as the matrix spike.   When the results  of  the  matrix spike
      analysis indicate a potential problem due to the sample matrix itself, the
      LCS  results   are  used  to  verify that  the  laboratory  can  perform  the
      analysis in a clean matrix.

            8.4.3   See  Method 8000,  Sec.  8.0 for the  details  on  carrying out
      sample quality control procedures for preparation and analysis.

      8.5   It  is   recommended that  the laboratory  adopt  additional  quality
assurance practices for use with  this method.   The  specific practices that are
most productive depend upon the needs  of the  laboratory  and the  nature  of the
samples.   Whenever  possible,  the  laboratory should analyze standard reference
materials and participate in relevant performance evaluation studies.
                                   8332 - 3                         Revision 0
                                                                  January 1995

-------
9.0 METHOD PERFORMANCE

      9.1   Method  8332  was  tested  by  one laboratory  using two  industrial
wastewater samples.  The wastewaters were analyzed at two dilutions with three
injections of each  dilution.  The results of these analyses are  shown in Table 1.


10.0  REFERENCES

1.    U.S. Department of the Treasury;  Bureau of Alcohol, Tobacco and Firearms;
      Lab Number 88-N-0648 B.
                                   8332 - 4                         Revision 0
                                                                  January 1995

-------
                                    TABLE  1

                          SINGLE LABORATORY  PRECISION
                             Nitroglycerine by Weight
                  As parts per Million          As Percent

                      1036 ± 38                 0.1036 ± 0.0038
                       952 ± 53                 0.0952 ± 0.0053
The numbers  shown are  the  average concentration  plus  or minus  one standard
deviation calculated for six analyses on each sample.
                                   8332 -  5                         Revision 0
                                                                  January 1995

-------
                        METHOD 8332

NITROGLYCERINE BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
              7.1 Establish
                appropriate
             chromatographic
                conditions.
              7.2 Perform
              calibration.
                   7.3
                  Does
             sample response1
                 exceed
               calibration
                 range?
           7.3 Calculate sample
              concentrations.
                         8332 - 6
  Revision  0
January  1995

-------
                                  APPENDIX A

      THIN-LAYER CHROMATOGRAPHY PROCEDURE FOR QUALITATIVE  DETERMINATION
                       OF THE PRESENCE OF NITROGLYCERINE
A. 1.0  APPARATUS

       A. 1.1   Thin-layer  chromatography (TLC)  system

            A.1.1.1   TLC plate, Analtech silica gel  GHL  (Catalog  #  11511),  or
       equivalent.

            A.1.1.2   Device  (syringe,  capillary  pipet or other)  for  spotting
       TLC plate.

            A.1.1.3   TLC developing tank.

            A.1.1.4   Sprayer.


A.2.0  REAGENTS

       A.2.1   Solvents

            A.2.1.1  Acetonitrile, CH3CN.

            A.2.1.2  1,2-Dichloroethane, C1CH2CH2C1.

            A.2.1.3  Carbon tetrachloride, CC14.

            A.2.1.4  Ethanol  (absolute), CH3CH2OH.

            A.2.1.5  Acetone, CH3COCH3.

            A.2.1.6  Tetrahydrofuran, C4H80.

       A.2.2  Diphenylamine,  (C6H5)2NH.

       A.2.3  alpha-Naphthylamine,  C10H7NH2.

       A.2.4  Sulfanilic acid, 4-(H2N)C6H4S03H.

       A.2.5  Sulfuric  acid,  H2S04.

       A.2.6  Potassium Hydroxide,  KOH.

       A.2.7  Acetic  acid  (30%),  CH3C02H.

       A.2.8    TLC  developing   solvent  -   20%   dichloroethane/80%  carbon
tetrachloride.

       A.2.9 TLC overspray - Prepare a solution of  5% diphenylamine in ethanol.

                                    8332-7                           Revision  0
                                  Appendix A                       January  1995

-------
For very weak spots, a solution of 5% diphenylamine in concentrated sulfuric acid
may be required.

       A.2.10  Alternative TLC overspray (Greiss  reagent)  -
            1.  Alcoholic KOH (2%)
            2.  1% sulfanilic acid in acetic acid (30%)
            3.  1% alpha-naphthylamine in acetic acid (30%)

       A.2.11  TLC standard - Use double-base smokeless powder (e.g. Hercules Red
Dot, Bullseye,  or  Winchester-Western  ball  powder) available at any  gun shop.
Dissolve in acetone, acetonitrile, or THF.


A.3.0  PROCEDURE

       A.3.1   Add a quantity of the developing solvent mixture to the developing
tank that will  be  sufficient to continuously  wet the edge of the  TCL plate.
Cover the tank tightly, and allow  the vapor phase to equilibrate (several hours).

       A.3.2   Streak or  spot the  sample near  the edge of the plate,  but above
the area that will  be immersed in the developing solvent.  Allow streaked area
to dry completely.

       A.3.3   Streak or  spot the standard solution next to the sample spot, at
the same distance from the edge of the plate as the sample  spot.  Allow streaked
area to dry completely.

       A.3.4   Stand  the  plate in the  equilibrated  developing tank  and cover
tightly.  Let the plate develop until  capillary action has carried the solvent
nearly to the top of the plate.

NOTE:  It may be necessary  to adjust  the developing  time to  achieve suitable
       separation of the  nitroglycerine  from interfering  sample components.

       A.3.5   Remove the  plate from the  tank and  air  dry.

       A.3.6   Spray the developed plates with  a 5% solution of diphenylamine in
ethanol.  After spraying developed TLC plates, expose the plates  to U.V. light
(longwave, shortwave, or both simultaneously) for 15-30  minutes,  until spots
appear.  For very weak  spots, spray again with  5% diphenylamine in concentrated
sulfuric acid.

       A.3.7   Spray  developed plates with  Greiss reagent (Section A.2.10)  and
warm the plates in a 100°C oven for 5-10 minutes.

       A.3.8   The presence of a spot at  the same distance  from the origin as the
standard (Rf of approximately 0.4) indicates that  nitroglycerine may be present.
The absence  of  a spot in  this location indicates that  nitroglycerine is  not
present  above the detection  limit of the test.
                                    8332-8                          Revision 0
                                  Appendix A                      January 1995

-------
4.3  DETERMINATION OF ORGANIC ANALYTES

     4.3.4  INFRARED METHODS

         The following methods are included in this section:

         Method 8410:      Gas  Chromatography/Fourier  Transform  Infrared
                           (GC/FT-IR)    Spectrometry    for    Semi volatile
                           Organics:  Capillary Column
         Method 8430:      Analysis  of Bis(2-chloroethyl)ether  Hydrolysis
                           Products by Direct Aqueous Injection GC/FT-IR
         Method 8440:      Total  Recoverable  Petroleum   Hydrocarbons  by
                           Infrared Spectrophotometry
                                 FOUR  -  13                        Revision  3
                                                               January  1995

-------
                                  METHOD 8430

         ANALYSIS OF BIS(2-CHLOROETHYL) ETHER AND HYDROLYSIS PRODUCTS
                     BY DIRECT AQUEOUS  INJECTION GC/FT-IR
1.0  SCOPE AND APPLICATION

      1.1  This method provides procedures for the identification and
quantitation of bis(2-chloroethyl) ether and its hydrolysis compounds in
aqueous matrices by direct aqueous injection (DAI) and gas chromatography with
detection by a fourier transform infrared spectrometer (GC/FT-IR).  The
following compounds can be determined by this method:
Compound Name                        Abbreviation           CAS Number3
Bis(2-chloroethyl) ether
2-Chloroethanol
2-(2-Chloroethoxy)ethanol
Diethylene glycol
Ethyl ene glycol
BCEE
CE
2CEE
DEG
EG
111-44-4
107-07-3
628-89-7
111-46-6
107-21-1
   a  Chemical  Abstract Services Registry Number.

      1.2  Although the initial study upon which this method is based targeted
only the bis(2-chloroethyl) ether and its hydrolysis compounds, its has been
suggested that this method can be used for the identification of compounds
that are generally non-extractable, highly water soluble, thermally stable,
and do not co-elute with water from the GC.  Possible analytes include ethers
and alcohols.

      1.3  The minimum identifiable quantities (MIQ) for the five compounds in
organic-free reagent water range from a low of 46 ng for BCEE, to a high of
120 ng for EG (See Sec. 9.2 for MIQ definition).  The MIQ for a specific
sample may differ depending on the nature of the interferences in the sample
matrix and the amount of sample used for the analysis.  The method detection
limit (MDL) has not yet been determined for the target analytes.

      1.4  This method is restricted to use by or under the supervision of
analysts experienced in the use of gas chromatography (GC), the interpretation
of FT-IR spectra and the use of continuous data collection systems.  Each
analyst must demonstrate the ability to generate acceptable results with this
method.
2.0  SUMMARY OF METHOD

      2.1  Water samples are filtered through a 0.45 jum filter, and 1 /iL
aliquots are injected directly into a GC.  The GC is equipped with 2


                                   8430 - 1                         Revision 0
                                                                  January 1995

-------
detectors, a thermal conductivity detector (TCD) and an FT-IR.  During
analysis, the analyst disconnects the FT-IR from the system to prevent aqueous
degradation of the KBr window, using an 8 port switching valve.  Further
analysis of either the solid precipitate from the filtering step or the
aqueous filtrate for trace amounts of non-water soluble compounds may be done
by extracting the samples using appropriate 3500 series methods.


3.0  INTERFERENCES

      3.1  Method interference may be caused by contaminants in solvents,
reagents, glassware and other sample processing hardware that lead to discrete
artifacts and/or elevated baselines in the chromatograms.   All of these
materials must be routinely demonstrated to be free from interferences under
the conditions of the analysis by running laboratory method blanks.

             3.1.1  Glassware must be scrupulously cleaned.  Clean all
      glassware as soon as possible after use by rinsing with the last solvent
      used.  This should be followed by detergent washing with hot water, and
      rinses with tap water and organic-free reagent water.  It should then be
      drained dry, and heated in a laboratory oven at 130°C for several hours
      before use.  Solvent rinsing with methanol may be substituted for the
      oven heating.  After drying and cooling, glassware should be stored in a
      clean environment to prevent any accumulation of dust or other
      contaminants.

             3.1.2  The use of high purity reagents and solvents helps to
      minimize interference problems.  Purification of solvents by
      distillation in all-glass systems may be required.

      3.2  Matrix interferences may be caused by contaminants that are in the
sample.  The extent of matrix interferences will vary considerably from source
to source, depending upon the nature and diversity of the matrix being
sampled.  If significant interferences occur in subsequent samples, additional
cleanup may be necessary.

      3.3  The extent of interferences that may be encountered in this method
has not been fully assessed.  Although the GC conditions described allow for a
unique resolution of the specific compounds covered by this method, other
matrix components may interfere.


4.0  APPARATUS AND MATERIALS

      4.1  GC/FT-IR System

             4.1.1  GC - Temperature programmable oven equipped with a cool
      on-column injection port, a purged splitless injection port, or an
      equivalent suitable for capillary glass columns.
             4.1.2  Column  - 30 m DB-wax, 1.0 p,m  film, Megabore  (J&W
      Scientific), or equivalent.
                                   8430 - 2                         Revision 0
                                                                  January 1995

-------
             4.1.3  Detectors

                      4.1.3.1  Thermal  Conductivity Detector (TCD)  -  Must be
             able to  handle  temperatures up to 300°C.

                      4.1.3.2  FT-IR Spectrometer - System should be  equipped
             with a mercury-cadmium-telluride detector, using a  light-pipe
             interface with  KBr windows (available from Digilab  Model GC/C32
             or equivalent).   Resolution of 8 cm"1  and  a range of 4,000  to
             7,000 cm"1 is  required.  The light-pipe  interface and  transfer
             lines should  be contained in a heated system  (up to 250°C)  to
             prevent  sample  condensation.  Extra transfer  lines  will be
             needed for the  switching system.

             4.1.4  Detector Switching System - 8  port stainless steel GC
      rotary switching valve, with an inert interval surface (such as FEP
      Teflon) and capable of withstanding temperatures up to 300°C.  Ideally,
      the valve should be mounted inside the GC oven with external  control.
      If internal mounting is not possible then use of an external  heated
      valve enclosure may be employed.   This allows direct injection of
      aqueous samples by using the valve system to route water away from the
      FT-IR KBr window which water rapidly destroys.

             4.1.5  Data Collection  - Each detector  should have  its own
      signal recorder.

                      4.1.5.1  TCD Signal  -  Either  an analog  strip  chart
             recorder or a digital computerized data collection  system  is
             acceptable.

                      4.1.5.2  FT-IR Signal  - The continuous  collection  of the
             spectrometer's  signal will require a computerized data system
             with the ability  to  continuously collect spectra at a rate  of 4
             scans/sec and add the 4 scans to produce a data point for each
             second.  The  ability to compare the required  spectra to a
             library  of spectra may  also be useful  for confirmation purposes.

      4.2  Glassware

             4.2.1  Glass  Funnels

             4.2.2  Volumetric Flasks (square)  - various sizes.

             4.2.3  Pipettes (A grade) - various sizes.

             4.2.4  Sample Vial with teflon lined lids.

      4.3  Syringes -  10 ^l, suitable for GC work.

      4.4  Analytical  Balance,  accurate to ±0.0001  g.
      4.5  Vacuum Filtration Apparatus  -  0.45 jttm filter and clean glass
flasks that are able to hold at least 1 L  of liquid.
                                   8430 - 3                         Revision 0
                                                                  January 1995

-------
5.0  REAGENTS

      5.1  Reagent grade chemicals shall be used in all  tests.  Unless
otherwise indicated, it is intended that all reagents shall  conform to the
specifications of the Committee on Analytical  Reagents of the American
Chemical Society, where such specifications are available.  Other grades may be
used, provided it is first ascertained that the reagent  is  of sufficiently
high purity to permit its use without lessening the accuracy of the
determination.

      5.2  Organic-free Reagent Water - All references to water in this method
refer to organic-free reagent water,  as defined in Chapter  One.

      5.3  Helium Gas - Suitable for gas chromatography.

      5.4  Methanol, CH3OH -  Pesticide reagent  grade,  nanograde,  or
equivalent.

      5.5  Stock Standard Solutions

              5.5.1  Prepare, in organic-free reagent water, a stock standard
      solution containing all of the target analytes at  a concentration of
      1000 mg/L.  Record the actual weight of each compound added and
      calculate the actual concentration of each component  of the solution.

              5.5.2  Transfer the stock  standard solution into a Teflon-sealed
      screw-cap bottle for storage.  Store at 4°C and protect from light.
      Check the solution periodically for signs of degradation or evaporation.
      This solution must be replaced every 3 months, or  when any sign of
      degradation or evaporation is observed.

      5.6  Calibration Standards

              5.6.1  Calibration standards at 500, 250, 100, 50, and 25 mg/L,
      from the aqueous stock standard solution, by appropriate volumetric
      dilutions with water are suggested.  Store as in Sec.  5.5.2.  The
      calibration solutions should not be made by serial  dilution of a single
      solution.

              5.6.2  Since the MDL for the target compounds in water has not
      been established, the suggested calibration curve  concentrations may be
      modified, depending on matrix interferences and sensitivity of
      equipment.  Generally speaking, the calibration curve should span at
      least one order of magnitude and the working range should bracket the
      analyte concentration.

      5.7  Spiking Solution - The analyst should monitor the performance of
the analytical system and the effectiveness of the method in dealing with each
sample matrix, by spiking a known amount of the target analytes into blanks or
into matrix spike samples.  A suggested preliminary concentration of the
spiking stock solution is 100 g/L.  When this method becomes better
established, the spiking concentration should be set at  1 to 5 times higher
than the background concentration determined for that matrix.


                                   8430 - 4                         Revision 0
                                                                  January 1995

-------
      5.8  Surrogate Spiking Solution:  To monitor the performance of the
method for all samples, a minimum of one surrogate compound should chosen by
the laboratory.  This compound should be diluted with water to an appropriate
concentration and added to all samples, method blanks, matrix spikes, and
calibration standards.
6.0  SAMPLE COLLECTION, PRESERVATION AND HANDLING
      6.1  See the introductory material to this chapter, Organic Analytes,
Sec. 4.1.

      6.2  Samples should be stored at 4°C and protected from light.

      6.3  Samples should be filtered and analyzed within 14 days of
collection.
7.0  PROCEDURE

      7.1  GC/FT-IR Conditions (Recommended):

      Flow Rate:                   2.4 mL/minute of helium carrier gas.
      Run Time:                    approximately 15 minutes
      Injection Volume:            1 jitL
      Valve Switch Time:           4 minutes
      Temperature Program:         80"C to 220°C at 15°C/minute,  hold at 220°C
                                   for 10 minutes.
      TCD Temperature:             290°C
      GC/FT-IR Interface,
      Transfer Lines, and
      Light Pipe Temperature:      220°C
      Scan Time:                   4 scans/second

NOTE: Modifications of these parameters may be necessary to facilitate the
      separation of certain compounds depending on matrix interferences
      encountered.

      7.2  External Calibration

             7.2.1  Calibration standards may be prepared using the suggested
      concentrations in Sec. 5.6.   Matrix interferences may prevent
      quantitation at the suggested concentrations.  When necessary,  the
      lowest concentration of the  calibration curve should be adjusted to be
      at, or near, the method detection limit.  Refer to Method 8000, Sec. 7.0
      for proper external calibration procedures.

             7.2.2  Refer to Method 8000, Sec. 7.0 for the establishment of
      retention time windows.

             7.2.3  Analyze a solvent blank to ensure that the system is
      clean and interference free.
                                   8430 - 5                         Revision 0
                                                                  January 1995

-------
       7.2.4  Analyze the 5 calibration standards, starting with the
lowest concentration and ending with highest concentration.

       7.2.5  Tabulate the IR absorbance peak area along with the
calibration factor (CF) for the analyte at each concentration.  Refer to
Sec. 7.0 of Method 8000 for linear and non-linear calibration acceptance
criteria.  It should be noted that IR transmission is not directly
proportional to concentration.

       7.2.5  Recheck the instrument calibration each day, before and
after an analysis is performed, by analyzing one or more calibration
standards.  The response obtained should fall  within ±15 percent of the
expected value or the instrument must be recalibrated.

       7.2.6  After the analysis of 10 or fewer samples, one of the
calibration standards must be reanalyzed to ensure that the retention
times and the CFs of the target analytes remain within the QC
requirements.

7.3  Sample Spiking and Filtering

       7.3.1  Allow the sample to come to ambient temperature.  Mark
the water meniscus on the side of the 1 L sample bottle for later
determination of exact sample volume.

       7.3.2  Add 2 ml of the spiking solution to the spiked blank and
the matrix spike sample.  The final  concentration of the added analytes
should be about 200 mg/L.

       7.3.3  Vacuum filter the sample through a 0.45 /urn filter that
has been rinsed with organic-free reagent water.  The filtrate should be
collected in a clean glass bottle.  Any particulate collected by the
filtering process may be discarded or extracted for analysis of trace
amounts of non-water soluble compounds using an appropriate 3500 series
method.

       7.3.4  Using a pipette, withdraw a 5 ml aliquot of the sample
(aqueous filtrate) and place it into a clean glass sample vial with a
teflon lined lid.  This 5 ml aliquot will  be the sample used for the
direct aqueous injection of the water sample.   Store at 4°C in the dark.
The remainder of the aqueous filtrate may be saved in a glass bottle
with a teflon lined lid until after the analysis is complete or
extracted for analysis of trace amounts of non-water soluble compounds
using an appropriate 3500 series method.
7.4  GC Analysis

       7.4.1  Method 8000, Sec. 7.0 provides instructions on
calibration, establishing retention time windows, the analysis sequence,
appropriate dilutions, and identification criteria.


                             8430 - 6                         Revision 0
                                                            January 1995

-------
             7.4.2  Determining the Valve Switching Time - The valve switch
      time must be determined before proceeding with the analysis of samples.

                      7.4.2.1   Place  the  GC  switching  valve  in  the  position
             which routes the sample to the TCD and away from the FT-IR
             spectrometer and inject a 1 ^L aliquot of  a reagent blank to
             determine the retention time of water in the system.  Use this
             retention time to determine the optimum valve switch time to
             cycle water away from the KBr window loop.  At the optimum valve
             switch time, the 8 port valve may be moved to the position which
             allows use of the FT-IR detector and the TCD in tandem after the
             bulk of the aqueous portion of the sample  has been diverted away
             from the KBr window (See Figure 1).

                      7.4.2.2   Once the optimum  valve  switch time is
             established, the system may be re-evaluated with one of the
             aqueous calibration standards to assure complete separation of
             all target analytes.

NOTE: Traces of water in the GC/FT-IR interface are acceptable.  However,  the
      repeated injection of 1.0 jiL aqueous samples may  eventually destroy the
      KBr window used for FT-IR detection.   Care should be taken in deciding
      the valve switching time so that the KBr window is exposed to only trace
      amounts of water and target analytes that elute with retention times
      just after water are not missed.

                      7.4.2.3   If the  TCD  is proven  to  have  adequate
             sensitivity for a particular analysis, it may be used as the
             prime detector once the target analytes are identified using the
             FT-IR detector.  When the TCD is used as the prime detector,  the
             switching valve should remain in the position which diverts the
             sample away from the KBr window.

             7.4.3  Sample Analysis - Analyze the samples, blanks,  spiked
      blanks, and spiked matrix samples  by injecting 1 juL aliquots into the
      GC and switching to the FT-IR spectrometer at the previously  determined
      time.   Dilution of the sample may  be necessary to adjust  the  analyte
      concentration to within the working range of the calibration  curve.   If
      dilution is necessary,  note which  samples were diluted in the final
      report and make the appropriate calculation  adjustments.

             7.4.4  GC/FT-IR Identification -  Visually compare the analyte
      infrared (IR)  spectrum versus the  search library spectrum of  the  most
      promising on-line library search hits.   Report,  as identified,  those
      analytes with IR frequencies  for the five (maximum number)  most intense
      IR bands (S/N > 5)  which are  within ±5.0 cm"1 of the corresponding bands
      in the library spectrum.   Choose IR bands which  are sharp and well
      resolved.  The software used  to locate spectral  peaks  should  employ  the
      peak "center of gravity" technique.   In  addition,  the  IR  frequencies of
      the analyte and library spectra should be determined with the same
      computer software.
                                   8430 - 7                         Revision 0
                                                                  January 1995

-------
      7.5  Calculations

             7.5.1  Calculate the calibration factor for each calibration
      standard and determine the percent relative standard deviation (%RSD)
      using the external standard calibration procedure in Sec.  7.0 of Mehtod
      8000.

             7.5.2  Calculation of the concentration of analytes using the
      external  standard calibration procedure is provided in Sec. 7.0 of
      Method 8000.
8.0  QUALITY CONTROL

      8.1  Refer to Chapter One and Method 8000 for specific quality control
procedures.  Quality control procedures to validate sample extraction is
covered in Method 3500.  Each laboratory should maintain a formal  quality
assurance program.  The laboratory should also maintain records to document
the quality of the data generated.

      8.2  Quality control necessary to evaluate the GC system operation is
found in Method 8000, Sec. 7.0 under Retention Time Windows, Calibration
Verification and Chromatographic Analysis of Samples.   Refer to Appendix A for
FT-IR spectrometer QC requirements.

      8.3  Initial Demonstration of Proficiency - Each laboratory must
demonstrate initial proficiency with each sample preparation and determinative
method combination it utilizes, by generating data of acceptable accuracy and
precision for target analytes in a clean matrix.  The laboratory must also
repeat the following operations whenever new staff are trained or significant
changes in instrumentation are made.  See Method 8000, Sec. 8.0 for
information on how to accomplish this demonstration.

      8.4  Sample Quality Control for Preparation and Analysis - The
laboratory must also have procedures for documenting the effect of the matrix
on method performance (precision, accuracy, and detection limit).   At a
minimum, this includes a method blank, matrix spike, a duplicate,  a laboratory
control sample (LCS) in each analytical batch and the addition of surrogates
to each field sample and QC sample.

             8.4.1  Documenting the effect of the matrix should include the
      analysis of at least one matrix spike and one duplicate unspiked sample
      or one matrix spike/matrix spike duplicate pair.  The decision on
      whether to prepare and analyze duplicate samples or a matrix
      spike/matrix spike duplicate must be based on a knowledge of the samples
      in the sample batch.  If samples are expected to contain target
      analytes, then laboratories may use one matrix spike and a duplicate
      analysis of an unspiked field sample.  If samples are not expected to
      contain target analytes, the laboratories should use a matrix spike and
      matrix spike duplicate pair.

             8.4.2  A Laboratory Control Sample (LCS) should be included with
      each analytical batch.  The LCS consists of an aliquot of a clean
      (control) matrix similar to the sample matrix and of the same weight or

                                   8430 - 8                         Revision  0
                                                                  January 1995

-------
      volume.  The LCS is spiked with the same analytes at the same
      concentrations as the matrix spike.  When the results of the matrix
      spike analysis indicates a potential problem due to the sample matrix
      itself, the LCS results are used to verify that the laboratory can
      perform the analysis in a clean matrix.

             8.4.3  See Method 8000, Sec. 8.0 for the details on carrying out
      sample quality control for preparation and analysis.

      8.5  Surrogate recoveries:     The  laboratory should evaluate surrogate
recovery data from individual samples versus the surrogate control  limits
developed by the laboratory.  Currently, surrogate compounds have not been
selected for this procedure.  See Method 8000, Sec. 8.0 for information on
evaluating surrogate data and developing and updating surrogate limits.

      8.6  It is recommended that the laboratory adopt additional quality
assurance practices for use with this method.  The specific practices that are
most productive depend upon the needs of the laboratory and the nature of the
samples.  Whenever possible, the laboratory should analyze standard reference
materials and participate in relevant performance evaluation studies.


9.0  METHOD PERFORMANCE

      9.1  No MDL data are available.

      9.2  Minimum Identifiable Quantities (MIQ)

             9.2.1  The MIQ  is defined as the minimum quantity that must be
      injected to result in a spectral  match that has the correct compound
      identification in the top 5 spectral matches.  The MIQ will vary
      depending on instrument sensitivity and sample matrix effects.

             9.2.2  The MIQ  range for CE, BCEE, EG, 2CEE, and DEG in organic-
      free reagent water by direct aqueous injection goes from a low of 46 ng
      for BCEE to a high of 120 ng for EG.1


10.0  REFERENCES

1.    Payne,  W.D. and Collette,  T.W., "Identification of Bis(2-chloroethyl)
      ether Hydrolysis Products by Direct Aqueous Injection GC/FT-IR," J.  of
      High Res.  Chrom..  12,  693-696,  1989.
                                   8430 - 9                         Revision 0
                                                                  January 1995

-------
                                                METHOD  8430
               ANALYSIS OF BIS(2-CHLOROETHYL)  ETHER  AND HYDROLYSIS  PRODUCTS
                                      BY DIRECT  AQUEOUS  INJECTION
    7.1 Set operating con-
   ditions of the GC/FT-IR
   to facilitate seperation
       of compounds
   7.2 Perform calibration
   (calibration should be
  checked each day and a
 calibration standard  should
  be frequently analyzed to
 ensure  retention times and
   RFs remain within QC
      requirements.)
7.3.1 Allow sample to come
 to ambient temperature and
mark the water meniscus on
 the sample bottle for later
   volume determination.
   7.3.2 Add 2 ml of the
   spiking solution to the
    spiked bland and the
    matrix spike sample.
    7 .3.3  Filter sample
    through a 0.45 /im
          filter.
   7.3.4 Withdraw a bmL
   aliquot and store at 4°C
   in the dark (to be used
for DAI of the water  sample.)
7.4.1  Refer to Method 8000
 for instructions on analysis
   sequence, appropriate
dilutions, and mdentification
         criteria.
  7.4.2 Establish optimum
  valve switch time and re-
   evaluate the system to
 ensure complete separation
   of the target analysis.
  7.4.3 Establish optional
   retention time window
  (refer to Method 8000.)
                                                  7.4.4 Inject samples,
                                                  blanks, spiked blanks,
                                                and spiked matrix samples
                                                      and analyze.
           7.4.4
         Is analyte
       concentration
     within the working
        range of the
        calibration'
7.4.4 Dilute and
   reanalyze.
 745 Visually compare the
   analyte IR spectrum to
the search library spectrum.
 7.4.5 Report as identified,
   those analytes with IR
    frequencies within
  + /-   5.0 cml for the five
  (maximum number) most
intense corresponding bands
  in the library  spectrum.
                                                 7.5.1 Calculate each
                                                 response factor using
                                                  equations provided.
       7.5.2 Calculate the
      concentrations of the
      analytes (see Method
            8000.)
                                               8430  -  10
                                                                                            Revision  0
                                                                                         January  1995

-------
                                  APPENDIX A
                     Quality Control  of FT-IR  Spectrometer

A.I  Equipment Adjustments and Maintenance

      A.1.1  Mirror Alignment - Adjust the interferometer mirrors ti attain
the most intense signal.  Data collection should not be initiated until the
interferogram is stable.  If necessary, align the mirrors prior to each GC/FT-
IR run.

      A.1.2  Interferometer - If the interferometer is air-driven, adjust the
interferometer drive air pressure to manufacturer's specifications.

      A.1.3  Lightpipe - The lightpipe and lightpipe windows should be
protected from moisture and other corrosive substances at all times.  For this
purpose, maintain the lightpipe temperature above the maximum GC program
temperature but below its thermal degradation limit.  When not in use,
maintain the lightpipe temperature slightly above ambient.   At all times
maintain a flow of dry, inert, carrier gas through the lightpipe.

      A.1.4  Beamsplitter - If the spectrometer is thermostated, maintain the
beamsplitter at a temperature slightly above ambient at all  times.  If the
spectrometer is not thermostated, minimize exposure of the beamsplitter to
atmospheric water vapor.

A.2  Centerburst Intensity and MCT Detector Check

      A.2.1  With an oscilloscope, check the MCT detector centerburst
intensity versus the manufacturer's specifications.  Increase the source
voltage, if necessary, to meet these specifications.  For reference purposes,
laboratories should prepare a plot of time versus detector voltage over at
least a 5 day period.

      A.2.2  If the centerburst intensity is 75 percent or less of the mean
intensity of the plot maximum obtained by the procedure,  install a new source
and check the MCT centerburst with an oscilloscope versus the manufacturer's
specifications (if available).  Allow at least five hours of new source
operation before data acquisition.

      A.2.3  Align Test - With the lightpipe and MCT detector at thermal
equilibrium, check the intensity of the centerburst versus  the signal
temperature calibration curve.  Signal  intensity deviation  from the predicted
intensity may mean thermal  equilibrium has not yet been achieved, loss of
detector coolant,  decrease in source output,  or a loss in signal throughput
resulting from lightpipe deterioration.

A.3  GC/FT-IR Sensitivity

      A.3.1  Capillary Column Interface Sensitivity Test  -  Install a 30 m x
0.32 mm fused silica capillary column coated with 1.0 /Ltm of DB-5  (or
equivalent).  Set the lightpipe and transfer lines at 280°C,  the injector at
225°C and the GC detector at 280°C (if used).   Under splitless Grob-type or
on-column injection conditions,  inject 25 ng of nitrobenzene, dissolved in 1


                                   8430  -  11                         Revision 0
                                                                  January 1995

-------
fj.L of methylene chloride.  The nitrobenzene should be identified by the on-
line library software search within the first five hits (nitrobenzene should
be contained within the search library).

      A.3.2  One Hundred Percent Line Test - Set the GC/FT-IR operating
conditions to those employed for the Sensitivity Test (see Sec. A.3.1).
Collect 16 scans over the entire detector spectral range.   Plot the test and
measure the peak-to-peak noise between 1800 and 2000 cm"1.   This noise should
be less than or equal to 0.15 percent.  Store this plot for future reference.

      A.3.3  If the GC/FT-IR was purchased before 1985,  there may be a
temperature effect at the interface.  To account for this, prepare a plot of
lightpipe temperature versus MCT centerburst intensity (in volts or other
vertical height units).  This plot should span the temperature range between
ambient and the lightpipe thermal limit in increments of about 20°C.  Use this
plot for daily QA/QC (see Sec. A.2.3).  Note that modern GC/FT-IR interfaces
(1985 and later) may have eliminated most of this temperature effect.

A.4  Frequency Calibration - At the present time, no consensus exists within
the spectroscopic community on a suitable frequency reference standard for
vapor-phase FT-IR.  One reviewer has suggested the use of indene as an on-the-
fly standard.

A.5  Single Beam Test - With the GC/FT-IR at analysis conditions, collect 16
scans in the single beam mode.  Plot the co-added file and compare with a
subsequent file acquired in the same fashion several minutes later.  Note if
the spectrometer is at purge equilibrium.  Also check the plot for signs of
deterioration of the lightpipe potassium bromide windows.   Store this plot for
future reference.
                                   8430 -  12                        Revision 0
                                                                  January 1995

-------
                                 METHOD  8440

   TOTAL RECOVERABLE PETROLEUM HYDROCARBONS BY INFRARED SPECTROPHOTOMETRY



1.0  SCOPE AND APPLICATION

      1.1  Method 8440 (formerly Method 9073) is used for the measurement of
total recoverable petroleum hydrocarbons (TRPHs) extracted with supercritical
carbon dioxide from sediment, soil and  sludge samples using Method 3560.

      1.2  Method 8440 is  not applicable  to  the measurement  of gasoline and
other volatile petroleum fractions.

      1.3   Method 8440  can detect TRPHs  at concentrations  of 10 mg/L  in
extracts.  This translates  to 10 mg/Kg in soils when  a 3 g sample is extracted
by SFE  (assuming  100  percent extraction efficiency), and the  final  extract
volume is 3 mL.

      1.4  This method  is  restricted  to use by or  under  the  supervision  of
trained  analysts.   Each  analyst must  demonstrate  the  ability to  generate
acceptable results with this method.


2.0  SUMMARY OF METHOD

      2.1  Soil samples are extracted with supercritical carbon dioxide using
Method 3560.   Interferences are removed with silica gel, either by shaking the
extract with loose silica  gel, or  by passing  it through  a silica  gel  solid-
phase extraction  cartridge.  After infrared  (IR)  analysis  of  the  extract,
TRPHs are quantitated by direct comparison with standards.


3.0  INTERFERENCES

      3.1  The analyte  class being measured  (TRPHs)  is defined within  the
context of this method.   The measurement may be subject to interferences,  and
the results should be interpreted accordingly.

      3.2  Determination  of TRPHs is a measure of mineral oils only,  and does
not include the biodegradable animal greases  and  vegetable oils captured  in
oil and  grease  measurements.   These non-mineral-oil contaminants may  cause
positive interferences with IR analysis,  if they  are not  completely  removed
by the silica gel  cleanup.

      3.3  Method  8440 is not appropriate for use in the analysis of gasoline
and other  volatile petroleum  fractions because  these  fractions  evaporate
during sample preparation.
                                  8440 -  1                         Revision 0
                                                                January  1995

-------
4.0  APPARATUS AND MATERIALS

      4.1   Infrared  spectrophotometer -  Scanning  or fixed wavelength,  for
measurement around 2950 cm"1.

      4.2  IR cells - 10 mm,  50 mm, and 100 mm pathlength, sodium chloride or
IR-grade glass.

      4.3  Magnetic stirrer with Teflon-coated stirring bars.

      4.4   Optional  - A vacuum  manifold  consisting of glass  vacuum basin,
collection  rack  and  funnel,  collection  vials,  replaceable stainless  steel
delivery tips, built-in vacuum bleed  valve and gauge  is  recommended for use
when silica gel  cartridges  are used.  The system is  connected to a vacuum pump
or water  aspirator through a vacuum  trap made from a 500 mL  sidearm  flask
fitted with a one-hole stopper and glass tubing.
5.0  REAGENTS

      5.1   Reagent-grade  chemicals  shall  be used  in all  tests.    Unless
otherwise indicated,  it  is intended  that all reagents shall  conform  to the
specifications  of  the  Committee  on  Analytical  Reagents  of the  American
Chemical Society, where such specifications are available.  Other grades may
be used, provided it is  first  ascertained  that the reagent is of sufficiently
high  purity  to  permit  its  use  without  lessening   the   accuracy  of  the
determination.

      5.2  Organic-free  reagent water.  All references  to water in this method
refer to organic-free reagent water, as defined in Chapter  One.

      5.3    Hydrochloric  acid   (1:1  v/v),   HC1.    Mix  equal  volumes  of
concentrated HC1 and water.

      5.4     Tetrachloroethylene,   C2C14   -   spectrophotometric   grade,  or
equivalent.

      5.5  Raw materials for reference oil  mixture - spectrophotometric grade,
or equivalent.

            5.5.1   n-Hexadecane, CH3(CH2)14CH3

            5.5.2   Isooctane, (CH3)3CCH2CH(CH3)2

            5.5.3   Chlorobenzene, C8H5C1

      5.6  Sodium sulfate (granular,  anhydrous),  Na2S04. Purify by heating at
400°C for 4 hours in a shallow tray, or by precleaning  the sodium sulfate with
methylene  chloride  (CH2C12).    If  the   sodium  sulfate  is  precleaned  with
                                  8440 - 2                        Revision 0
                                                                January 1995

-------
methylene chloride, a method blank must be analyzed, demonstrating that there
is no interference from the sodium sulfate.

      5.7  Silica gel.

            5.7.1    Silica  gel  solid-phase extraction  cartridges  (40 /urn
      particles, 60 A pores), 0.5 g, Supelco, J.T. Baker, or equivalent.

            5.7.2    Silica  gel,  60  to  200 mesh,  Davidson  Grade  950 or
      equivalent (deactivated with 1 to 2 percent water).

      5.8  Calibration mixtures:

            5.8.1  The material of interest, if  available, or  the  same  type
      of  petroleum  fraction,  if  it  is  known  and   original   sample is
      unavailable, shall  be used  for preparation of calibration  standards.
      Reference oil is to  be used only for unknowns.  Whenever possible,  a GC
      fingerprint  should  be  run on  unknowns  to determine  the  petroleum
      fraction type.

            5.8.2   Reference  oil  -  Pipet  15.0  ml n-hexadecane,  15.0 ml
      isooctane,  and 10.0 ml  chlorobenzene into  a 50  ml  glass-stoppered
      bottle.   Maintain  the integrity of  the  mixture  by keeping  stoppered
      except when withdrawing aliquots.   Refrigerate at 4°C when  not in  use.

            5.8.3  Stock standard - Pipet 0.5 ml calibration standard (Section
      5.8.1  or 5.8.2)  into a  tared  100  mL volumetric  flask and  stopper
      immediately.  Weigh and dilute to volume with tetrachloroethylene.

            5.8.4  Working  standards -  Pipet  appropriate volumes  of  stock
      standard  (Sec.   5.8.3)  into  100  ml volumetric flasks according  to the
      cell  size to  be  used.    Dilute to  volume with  tetrachloroethylene.
      Calculate the concentrations of the  standards from  the  stock standard
      concentrations.

      5.8   Calibration mixture  for  silica  gel  cleanup  -  Prepare a  stock
solution of corn oil  by placing  about  1  ml (0.5 to 1 g) of corn oil  into  a
tared 100 ml volumetric flask.   Stopper  the flask and weigh to the nearest
milligram.   Dilute  to  the  mark with  tetrachloroethylene,   and  shake the
contents to effect dissolution.   Prepare additional  dilutions to  cover the
range of interest.


6.0  SAMPLE COLLECTION,  PRESERVATION,  AND HANDLING

      6.1  Solid samples  should  be  collected and stored as any other solid
sample containing  semivolatile  analytes.   See  the introductory material to
this Chapter, Organic Analytes, Sec.  4.1.
                                  8440  -  3                        Revision 0
                                                               January  1995

-------
      6.2  Samples should be  analyzed with minimum delay, upon receipt in the
laboratory, and must be kept refrigerated prior to analysis.


7.0  PROCEDURE

      7.1  Prepare samples according to Method 3560.

      7.2  Add 0.3 g of loose silica gel to the extract and shake the mixture
for 5 minutes,  or pass the  extract through  a  0.5 g  silica gel  solid-phase
extraction cartridge  (conditioned  with  5  ml of tetrachloroethylene).   When
using loose  silica gel,  filter the extract  through a  plug  of  precleaned
silanized glass wool in a disposable glass pipette.

      7.3   After  the  silica gel  cleanup,  fill  a clean IR  cell  with  the
solution  and  determine the  absorbance  of the  extract.   If the  absorbance
exceeds the linear range of the IR spectrophotometer,  prepare  an appropriate
dilution and reanalyze.  The possibility that the  absorptive capacity of the
silica gel has  been exceeded can  be tested  at  this  point by repeating  the
cleanup and determinative steps.

      7.4   Select appropriate  working  standard concentrations  and  cell
pathlengths according to the following ranges:

                          Concentration range
Pathlength (mm)           fug/ml of extract)               Volume (ml)
     10                          5  to  500                        3
     50                          1  to  100                      15
    100                         0.5 to  50                      30

     Calibrate  the instrument for the  appropriate cells using  a series  of
working standards.   Determine absorbance  directly for each solution  at  the
absorbance  maximum  at  about  2950 cm"1.    Prepare  a  calibration  plot  of
absorbance versus concentration of  petroleum  hydrocarbons  in  the  working
standards.

     7.5  Determine the concentration of TRPHs in the extract by comparing the
response against  the calibration plot.

     7.6   Calculate  the concentration  of  TRPHs  in  the  sample  using  the
formula:
                                                R x D x  V
                    Concentration  (mg/Kg)  =    	
                                                    W
     where:

     R  =  mg/mL of TRPHs as  determined  from the calibration  plot
     V  =  volume of extract, in milliliters
     D  =  extract dilution  factor, if  used
     W  =  weight of solid sample,  in  kilograms.


                                  8440 - 4                        Revision 0
                                                                January 1995

-------
     7.7  Recover the tetrachloroethylene  used in this method by distillation
or other appropriate technique.


8.0  QUALITY CONTROL

     8.1   Reagent  blanks or matrix-spiked samples must be  subjected  to  the
same analytical procedures as those used with actual samples.

     8.2  Refer to Chapter One  for specific Quality Control procedures and to
Method 3500 for sample preparation procedures.
9.0  METHOD PERFORMANCE

     9.1  Table  1  presents a comparison of certified values  and  the  values
obtained using Methods 3560 and 8440.   Data are presented for both Freon-113
and tetrachloroethylene, since both solvents were found  to  be an  acceptable
collection solvent.   However,  only tetrachloroethylene  is  recommended  as  a
collection solvent for TRPHs in Method 3560.

     9.2   Table 2  presents precision  and accuracy data  from the  single-
laboratory evaluation  of  Methods  3560 and  8440 for  the   determination  of
petroleum hydrocarbons from spiked soil samples.  These  data  were obtained  by
extracting samples at 340 atm/80°C/60 minutes (dynamic).
10.0  REFERENCES

1.   Rohrbough, W. G.; et  al.   Reagent  Chemicals, American  Chemical  Society
     Specifications, 7th ed.; American Chemical  Society, Washington, DC, 1986.

2.   Methods  for  Chemical  Analysis  of Water and Wastes; U.S.  Environmental
     Protection Agency.  Office of  Research  and Development,  Environmental
     Monitoring and Support Laboratory.  ORD Publication Offices of Center for
     Environmental Research Information, Cincinnati,  OH,  1983;  EPA-600/4-79-
     020.
                                  8440 -  5                        Revision  0
                                                               January  1995

-------
                                  TABLE  1

               CERTIFIED AND SPIKE VALUES COMPARED TO RESULTS
                       OBTAINED BY METHODS 3560/8440
Reference Material
Environmental Resource Assoc.
TPH-1 (lot 91012)
Environmental Resource Assoc.
TPH-2 (lot 91012)
Clay spiked with keroseneb
Clay spiked with light gas oil0
Clay spiked with heavy gas oild
Environmental Resource Assoc.
TPH-1 (lot 91017)"
Environmental Resource Assoc.
TPH-2 (lot 91017)6
Spike cone, or
certified cone.
(mg/kg)
1,830
2,230
100
100
100
614
2,050
Methods
3560/8440
(mg/kg)
l,920±126a
2,150±380a
86.0; 93.0
84.0; 98.0
103; 108
562; 447
1,780; 1,780
8 Three 60  minute extractions.   The  extracted  material  was  collected  in
  Freon-113; the  concentrations  were determined  against  the reference  oil
  standard.

b Duplicate  30 minute extractions.  The extracted material was collected in
  tetrachloroethylene; the  concentrations  were determined against  standard
  made from  the spiking material.

0 Six  30  minute  extractions.   The  extracted  material  was  collected  in
  tetrachloroethylene; the concentrations  were determined  against  a  standard
  made from  the spiking material.

d Four 30  minute  extractions.    The extracted  material  was  collected  in
  tetrachloroethylene; the concentrations  were determined  against  a  standard
  made from  the spiking material.

6 Three 30  minute extractions.   The  extracted  material  was  collected  in
  tetrachloroethylene;  the  concentrations  were  determined  against   the
  reference  oil standard.
                                  8440 - 6
  Revision 0
January 1995

-------
                             TABLE 2

      SINGLE-LABORATORY METHOD ACCURACY AND PRECISION FOR
             METHODS 3560/8440 FOR SELECTED MATRICES
Spike cone, or Method
certified cone. Spike accuracy
Matrix (mg/kg) Material (% recovery)
Clay soil6
ERA TPH-18
(lot 91016)
ERA TPH-28
(lot 91016)
SRS103-100b
2,500 Motor oil 104
2,350 Vacuum oil 80.3

1,450 Vacuum oil 88.6

32,600 c 94.2
Method
precision
(% RSD)
8.5
19.7

19.6

4.0
Eight determinations were made using two different supercritical fluid
extraction systems.  The extracted material was collected  in Freon-113.

Ten determinations were made using three different supercritical fluid
extraction  systems.    The  extracted  material  was  collected  in
Freon-113.

This is a standard reference soil certified for polynuclear aromatic
hydrocarbons.  No spike was added.
                         8440 - 7
  Revision 0
January 1995

-------
                                METHOD  8440

 TOTAL RECOVERABLE PETROLEUM  HYDROCARBONS  BY  INFRARED SPECTROPHOTOMETRY
   Prepare sample(s)
    according to
appropriate Method.
                                                          7.5 - 7.6 Calculate
                                                         the concentration of
                                                         TPHs  in the sample.
                                                            7.7 Recover the
                                                          tetrachloroethylene
                                                             by appropriate
                                                               technique.
7.2 Proceed with
silica gel cleanup
 7.3 Determine the
   absorbance of
    the extract.
        7.3
       Does
absorbance excee
    linear range
     of the IR
     spectro-
    photometer
                               7.3 Dilute
                             appropriately
  7.4 Calibrate the
     instrument.
                                  8440  -  8
                                                                    Revision 0
                                                                  January 1995

-------
4.3 DETERMINATION OF ORGANIC ANALYTES

   4.3.5  MISCELLANEOUS SPECTROMETRIC METHODS

         The following method is included in this section:

         Method 8520:      Continuous Measurement of Formaldehyde in Ambient
                           Air
                                 FOUR  -  14                       Revision 3
                                                              January 1995

-------
                                  METHOD 8520

             CONTINUOUS MEASUREMENT OF FORMALDEHYDE  IN AMBIENT AIR
1.0  SCOPE AND APPLICATION

      1.1    This  method  is  applicable  to   the  continuous  measurement  of
formaldehyde (CAS No. 50-00-0)  in  ambient air.  This method is  for use primarily
for nonoccupational  exposure monitoring.

      1.2  This method is applicable to concentrations of formaldehyde from 6 to
500 p-g/m3.  Detection limits are dependent on sample airflow rate, with a maximum
rate set at 1.0 L/min.

      1.3   This method  is  restricted to  use  by  or under  the  supervision  of
trained  analysts.    Each  analyst must  demonstrate  the  ability  to  generate
acceptable results with this method.


2.0  SUMMARY OF METHOD

      2.1  A  modified portable commercial  analyzer which  consists of  a small
diaphragm pump,  an  analytical  module, a colorimeter with  glass  flow  cells,  a
peristaltic pump and tubing, and a 12  volt  rechargeable  battery for optional  DC
power  operation  is  used  for  this  method  (See  Figure  1).   An  acidified
pararosanil ine  scrubber  solution  is pumped through the  reference  cell  in  the
colorimeter and is then diluted 1:1  with  a  liquid flow stream containing water.
The diluted solution then passes into an air scrubber  coil where formaldehyde is
quantitatively absorbed from the air sample  stream.  After the liquid and air are
separated,  the sample solution  is mixed 1:1  with a dilute aqueous Na2S03 reagent.
A  series of  time delay  coils  allow  16  minutes  for  color  development  of  the
chromophore.   The colored product  is continuously detected  as  it passes through
the sample  cell by a UV-Visible spectrophotometer set at  550 nm and is recorded
on a strip  chart recorder or an automated data acquisition system.


3.0   INTERFERENCES

      3.1   Method  interferences  may  be caused  by contaminants  in  reagents,
glassware,  and sample processing hardware.

            3.1.1   Glassware must be  scrupulously cleaned.  Clean all glassware
      as soon  as possible after use by rinsing with  the last solvent used.  This
      should be followed by  detergent washing  with hot  water,  and  rinses with
      tap water and organic-free reagent  water.  It should  then be drained dry,
      and heated in  a laboratory  oven at  130°C for  several  hours  before use.
      Solvent  rinsing may be substituted for the oven  heating.  After drying and
      cooling, glassware should be stored in a clean environment,  covered with
      aluminum foil to prevent any accumulation of dust  or other contaminants.

            3.1.2   All  of  these materials  and equipment must be  routinely
      demonstrated to  be free  from interferences  under  the  conditions  of  the


                                   8520 - 1                         Revision  0
                                                                  January 1995

-------
      analysis by establishing the  analyzer  baseline with the introduction of a
      source of zero air (zero grade compressed air)  at the sample inlet.

            3.1.3  Instrument Cleaning

                  3.1.3.1  Reagent  Lines  - To ensure optimum performance of the
            analyzer, the three reagent  lines  should  be  placed  in a reservoir
            containing 1.0 M nitric acid as a  cleaning  solution.   Operate the
            analyzer with the  zero air supply and pump  the  cleaning solution
            through the analyzer for 1  hour.  To rinse the lines, pump organic-
            free reagent water  through the  system for 2 hours.   The cleaning
            sequence must be performed at frequent intervals (e.g. each second
            or third day of regular use).

                  3.1.3.2   Colorimeter Flow Cell  -   Clean  the  flow  cells  by
            periodically soaking  them  in  commercial  bleach  or  a  laboratory
            detergent solution.  After  cleaning the cells, rinse them well with
            organic-free reagent water.

      3.2   Although  there   is  some   disagreement  concerning  the  mechanism
involving  the  formation  of  the   final  product  in  the  reaction  between
pararosaniline and formaldehyde, it is  generally agreed that the chromophore is
an alkylsulfonic acid.  (Refs.  1,3)  Studies have  shown  that the color-forming
reaction  of  the  modified  pararosaniline  procedure   is highly  specific  to
formaldehyde.

            3.2.1      Low   molecular   weight   aldehydes   exhibit   positive
      interferences, but only when  present in  large excess  over formaldehyde.

            3.2.2  Sulfur dioxide,  at a concentration  of 520 M9/m3,  produces a
      slight negative interference.

      3.3   Sensitivity and  color development time  are  functions of temperature.
Good results may be obtained for operating temperatures between 15 and 35°C (with
optimum results at 25°C) by performing daily instrument calibrations.

      3.4   Baseline drift is negligible, or less  than ±2%  of full scale after
8 hours of continuous  operation,  under normal conditions.    Periodic baseline
checks should be performed to ensure accuracy.

      3.5   Low-concentration aqueous formaldehyde standard  solutions of 0.02 to
2.0 /jg/mL  and  dilute   sodium   sulfite  reagent solutions   exhibit  a  limited
stability and therefore must be prepared daily.

      3.6   The use  of high  purity reagents   and  solvents  helps to  minimize
interference problems.


4.0   APPARATUS AND MATERIALS

      4.1   Continuous Air Monitor
                                   8520 - 2                         Revision 0
                                                                  January 1995

-------
      4.1.1     Automated  Wet-chemical   Colorimetric  Analyzer   -  CEA
Instruments Model  TGM  555 air monitor equipped  with  AMK-4 formaldehyde
analytical module or equivalent.   The TGM 555 analyzer is equipped with a
dual beam colorimeter,  set  at  550 nm, consisting of glass reference and
sample flow cells,  a matched pair of photodetectors,  a miniature tungsten-
halogen lamp,  and a signal processing unit.  The colorimeter measures the
difference  in  light absorption  of the  pararosaniline chloride  reagent
before and after it reacts with formaldehyde.

      4.1.2  Spare  Parts - A spare parts maintenance kit is available from
CEA or the following equipment may be purchased.

            4.1.2.1  Silicon rubber peristaltic pump tubes (T-78) - Pump
      tubes should be replaced after 30  to 45 days of continuous use.

            4.1.2.2   Rubber diaphragm in the  miniature  air  pump  - The
      diaphragm should be replaced whenever  it  becomes worn or defective.

            4.1.2.3  Tungsten-halogen colorimeter lamp

      4.1.3  Glass  Coils (3 mm OD) and Connectors  -  These are required for
modifications of the analytical module to increase residence time for the
reactants.  The standard  version  of the AMK-4  analytical  module  (2 x 25
turn coil)  contains  a  residence  mixing  time of  less than  6  minutes, as
measured from the glass  double mixing tee to the entrance of the sample
cell. By  adding additional mixing  coils  (an   additional   120  turns) the
residence mixing time was  increased to 16 minutes  (absorbance reached a
maximum when the residence time of the reactants  was about 16 minutes or
longer).  These parts are used extensively in segmented flow colorimetric
analyzers such  as Technicon Instruments Corporations Autoanalyzer  II or an
equivalent.

      4.1.4  Glass  Debubbler  -  A second  glass debubbler  was  added just
ahead of  the  entrance  to  the  sample cell  to  overcome  the  increased
frictional drag on  the system.  The top exit  of  the debubbler was attached
to a T-78 silicon rubber tube on the peristaltic  pump so  that all of the
air between the sample segments is drawn  away to  waste and approximately
95% of the sample liquid is pushed into  the  sample cell  for measurement.
Further,  by  inverting  the glass  double  mixing  tee so  that  the sample
stream  enters  from the  top  and  the  Na2S03  reagent enters  from below,
smoother operation  (i.e., a more  regular bubble  pattern)  resulted.   All
connections from the exit of the glass double mixing tee  to the entrance
of the sample cell  are 3 mm OD glass tubing.

4.2   Zero Air Supply

      4.2.1   A regulated,  zero  grade  compressed air  cylinder  with
connections to  the  instrument sample inlet through a tee fitting to permit
atmospheric dumping of the excess air supply.

      4.2.2   Alternatively,  formaldehyde-free air may  be obtained  by
scrubbing  ambient  air  through  a  series  of  3  midget  impingers,  each
containing 15  ml of 0.05 M sulfuric acid,  followed by a silica-gel packed


                             8520 - 3                          Revision 0
                                                            January 1995

-------
      cartridge  for  drying the airstream.   The impinger  solutions  should be
      replaced daily.

NOTE: Avoid  the  use of  organic vapor  filters  which may  be supplied  with  a
      commercial instrument.    These  filters  have  proven to be unsuitable for
      use at formaldehyde concentrations less than 500
            4.2.3    Teflon  Tubing  and  Valves   -  FEP  Teflon  tubing  with  a
      0.25 in. OD and a  3-way  Teflon  valve  for  switching (manual  or solenoid)
      between the zero air supply and the ambient air sample stream.

      4.3   Data Acquisition System -  Data acquisition may be accomplished with
a manual strip chart recorder or an automated data acquisition system.

      4.4   Glassware

            4.4.1  Volumetric Flasks,  class A - 1 L,  250 ml, and other various
      sizes.

            4.4.2  Pipets, class A - various sizes

            4.4.3  Mohr  Pipets - 1.0 ml with 0.01 ml graduations

            4.4.4  Soap  Bubble Flowmeter  - 500 ml

      4.5   Thermometer  - to check the temperature at the time of sampling.


5.0   REAGENTS

      5.1   Reagent-grade chemicals shall  be used in all tests.  Unless otherwise
indicated, it is intended that  all  reagents  shall conform to the specifications
of the Committee on  Analytical  Reagents of the American Chemical Society, where
such specifications  are  available.   Other grades may be used,  provided  it is
first ascertained that the reagent  is  of  sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Organic-free Reagent Water -  All  references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Formaldehyde, CH20  -  ACS certified or assayed 36.5% solution (w/w),
store at room temperature and protect from light.

      5.4   Hydrochloric acid solution,  HC1  in water - 2.0 M

      5.5   Nitric Acid  solution, HN03 in water  - 1.0  N

      5.6   Sodium sulfite, Na2S03  - ACS  certified, anhydrous  crystal

      5.7   2.78 mM  Sodium sulfite  solution  -  Dissolve  0.175  anhydrous sodium
sulfite crystal  in 500 ml of water.  This solution should  be protected from light
during use and must  be prepared daily.
                                   8520 - 4                         Revision 0
                                                                  January 1995

-------
      5.8   Pararosaniline  chloride  (Basic  Fuchsin),  C20H20N3C1   -  specially
purified, 0.2% (w/v) in 1 M HC1 (Fisher Scientific, or equivalent).

      5.9   1.72 mM Pararosaniline chloride in 0.8 M HC1

            5.9.1   Using  a 100  ml graduated  cylinder,  measure  and  combine
      68.5 ml of the stock pararosaniline chloride solution  with 66 ml of 2.0 M
      HC1 solution in a 250 ml volumetric flask.

            5.9.2  Dilute the mixture with water to a  final  volume  of 250 ml.

            5.9.3  This solution should be protected from light during use.  The
      solution may be stored up to  3 months  if kept at 4"C and protected from
      light.

NOTE: The rate of  consumption  of  the 2.78 mM  sodium  sulfite  solution  and the
      1.72 mM pararosaniline chloride solution  is 0.6  L per  24  hours at routine
      analyzer operation rate of 0.40 mL/min.   After mixing  the three solutions
      together with  glass double mixing  tee,  the final  concentration  of the
      reactants should be  0.43 mM Pararosaniline, 0.2.  M HC1, and 1.39 mM Na2S03.

      5.10  Formaldehyde Stock Standard Solution - Add approximately 2.1  g of 37%
formaldehyde solution, weighed to  the nearest ±0.1 mg,  to a 1 L  volumetric flask
and dilute to  the mark to make a solution  of approximately  777 ng/ml.  Determine
the actual concentration of the solution  using  Method 8315.  The stock standard
formaldehyde solution  is  stable for up  to 1 year  if  it is  protected from the
light and stored at 4°C.

      5.11  Secondary Formaldehyde Standard Solutions - Prepare 2 more solutions
by  two  25:1 serial  dilutions  of the stock standard  solution with water  to
approximately 31.1 and  1.24 p.g/m\..    If  the  concentration of  the  formaldehyde
stock solution  is  not  777  /ug/ml_,  the actual  concentration of  the  secondary
standards must be calculated.  If  the solution  are kept in the  dark and at 4'C,
the 31.1 jig/mL standard may be stored for 3 months and the 1.24 jug/mL standard
may be stored for 7 days.

      5.12  Formaldehyde Calibration Standard Solutions -  Prepare 3 calibration
standards with concentrations  ranging from approximately 0.03 to  1.24  jug/mL.
Formaldehyde  concentration   in   the  above  range  are   equivalent   to   air
concentrations from 11.2 to 492 jug/m3. Working standards must be prepared daily.


6.0   SAMPLE COLLECTION

      6.1   Start-up Procedure

NOTE: The signal  response for a 100  ppbv  liquid standard remained unchanged for
      the temperatures  in the  20-30'C  range.    Check the  temperature  before
      analysis to  assure  environment to  be  analyzed is  within  the  proper
      temperature range.

            6.1.1  Activate the AC  power  switch  and the DC  power  switch which
      supplies power  to the colorimeter  lamp  and LED  digital display  panel.

                                   8520 - 5                         Revision 0
                                                                  January 1995

-------
Power  to  the  colorimeter  lamp  should  be  left  on  continuously  during
periods of regular use to maintain the thermal stability of the unit.

      6.1.2  Allow  a  minimum 2 hour warm-up  period  before  starting the
analyzer.   During periods of intermittent  regular use, keep the AC and DC
power switches on continuously, in order to minimize start-up time.

      6.1.3   Set the Range  switch  to the  "low"  position and  the  Damp
switch to the "high" position.

      6.1.4   Install  the 7 peristaltic pump tube  lines  in  their proper
sequence  on  the pump rollers.    Place  the  3  reagent  lines  in  a  water
reservoir, then switch the sample air inlet valve to the zero air supply
position.

      6.1.5  Activate the  Liquid  Flow  switch,  to  start  the  reagent  pump
and, at the same time, turn on the Air Flow switch to start  the air  pump
(using the zero air  supply).   Pump water for 20 to 30  minutes to allow the
system flows to stabilize and to verify trouble-free operation.

      6.1.6  Deactivate  the liquid flow pump  and the  airflow pump.  Place
each of the reagent  lines in its proper reagent reservoir and restart both
pumps.  Again,  using the  zero air supply,  allow  the reagents  to  pump
through the analyzer for 20 to 30 minutes.

      6.1.7  Follow  the operational manual  procedure for manual adjustment
of  the  optical  zero  on  the colorimeter  (if  necessary).   Use  the  zero
potentiometer to make fine adjustments for setting the baseline response
to a zero  setting on the  display  panel.  The span potentiometer should be
adjusted  to  the maximum   (10.0)  setting  for  most  low-concentration
formaldehyde measurements.

      6.1.8   Activate  the strip  chart   recorder  (or data  acquisition
system) and adjust  the baseline  response  to  read  5 to 10 percent of the
full scale response.  The selection of the  full-scale  voltage output range
depends  on  the  sensitivity  required  for  the  particular  monitoring
application.   A 1.0  volt  signal  response corresponds  to approximately
120 M9/m3 formaldehyde in the sample airstream.

      6.1.9  Monitor the baseline response of  the instrument for 10 to 15
minutes to verify that the signal is stable.

6.2   Routine Operating Procedure

      6.2.1  After the instrument baseline has stabilized using the  zero
air supply,  initiate the  multipoint  calibration  procedure  (as  in  Sec.
7.1.3) using the fresh liquid equivalent formaldehyde standard solutions
from Sec.  5.11.

      6.2.2  After sampling the final  liquid standard in the calibration
sequence,   return the reagent  line  that  is pumping  the  calibration
solutions to the water reservoir  and reestablish the  instrument baseline.
                             8520 - 6                         Revision 0
                                                            January 1995

-------
            6.2.3  Record  the  chart  speed  and  the exact starting time.  Begin
      monitoring the sample atmosphere by switching the sample air inlet valve
      from the zero air supply to the air sample position.  After a lag time of
      18 minutes, the  instrument  will  respond  quickly  (1  to  2 minutes) to the
      concentration of formaldehyde  in the airstream.

            6.2.4  Conduct  periodic baseline and calibration checks at intervals
      appropriate to the particular  operating characteristics of each analyzer
      system (see Sec. 7.1.3) and the specific monitoring application.

      6.3   Shutdown Procedure

            6.3.1  With the sample inlet valve in the zero air supply position
      and with the monitor in operation, place all 3 of the reagent lines in a
      reservoir containing water.

            6.3.2  Pump water through the instrument  for 30 minutes, then, shut
      off both the liquid  flow pump  and the airflow pump.   Remove  the 7 pump
      tubes from the  peristaltic pump rollers if the  instrument  is  to remain
      shut down for more than 3 hours.

            6.3.3  Deactivate the zero air  supply  (if necessary) and secure the
      reagent reservoirs for proper  storage.  Properly dispose of the contents
      of  the  waste  reservoir  in  accordance  with  acceptable  environmental
      procedures.

            6.3.4  Keep  the colorimeter lamp  on  unless the  instrument  is  to
      remain inactive for 7 days or more.

      6.4   Analysis of Data

            6.4.1  Determine the height of  each peak or plateau in the analysis
      data by measuring from the established baseline.

            6.4.2  Determine the  time  at which  any  peaks  or  plateaus occurred
      during the sampling analysis period.

            6.4.3  Using the  equation of the calibration  curve  determined  in
      Sec. 6.2.1, calculate  the concentration  of the formaldehyde  in  the  air
      sample,  (see Sec. 7.3.2)


7.0   PROCEDURE

      7.1   Calibration

            7.1.1  Calibration of the Air Pump

                  7.1.1.1   Using the guidelines  in the  instrument  operating
            manual,  adjust  the  stroke and electric  motor speed to  produce  a
            sample airflow rate of approximately 1.0 L/min. (Ref.  5)

                  7.1.1.2  Using a 500 ml soap bubble  flowmeter or other airflow
            rate calibration device and the airflow control potentiometer,  fine

                                   8520 - 7                         Revision  0
                                                                  January  1995

-------
            tune the airflow rate so that  it  is  calibrated  to  the nearest 1.0
            mL/min.

                  7.1.1.3  The  air  pump should be recalibrated every  30 days
            during regular operation,  whenever the pump diaphragm is replaced,
            or whenever there is a change in operational  airflow rate.

            7.1.2   Flow Rate  of the  Formaldehyde  Standard Peristaltic Pump
      Tubing - With analyzer operating  normally,  connect  a  1.0 ml Mohr pipet,
      with 0.01 mL graduations,  in line  with the formaldehyde solution delivery
      tube.  Calculate the  actual  liquid flow  rate, to the nearest 0.01  mL/min,
      by timing small  air bubbles introduced into the flow stream as they pass
      through  a  fixed  volume in the pipet.   Recalculate the  flow  rate  after
      every 7 days of regular operation or after replacement  of  the pump tubing.

            7.1.3  Colorimeter Calibration

                  7.1.3.1  Prepare  3  calibration standards which  bracket the
            working range of  the instrument  using  the concentration range given
            in Sec. 5.11.  The lowest concentration should be adjusted  to just
            above the method detection  limit.   Further,  the calibration  curve
            working range must bracket the sample analyte concentration.

                  7.1.3.2  Sample water for at least 10 minutes to establish a
            baseline and to be sure the system is clean and interference free.

                  7.1.3.3  Sample each  of the  3 standards for 10 to 15 minutes.
            Between each calibration standard a water sample should be  run for
            10 minutes.  The  calibration standards should be analyzed starting
            with the lowest concentration and ending with highest concentration.

                  7.1.3.4  Find the  established baseline from the analysis data
            and measure the height of each  of  the  3 calibration standard peaks.
            Plot  peak  height   versus   concentration   (/xg/m3  liquid  standard
            equivalent)  and   perform   a  linear   regression  analysis  of  the
            calibration data  to verify linearity.

                  7.1.3.5  A  full  3 point calibration should  be  performed at
            least once a week during routine  operation.   Analysis of a single
            mid-range calibration standard must be performed as a check at the
            beginning and end of any sampling period.  A mid-range calibration
            standard  must  also  be  analyzed  once  every  4 hours  to  verify
            instrument calibration.

NOTE: It  is  possible  to  calibrate  the  TGM  555  analyzer  using  gaseous
      formaldehyde  standards,   however,  the  procedures   required to  produce
      accurate,  dynamic, low-concentration  standard mixtures  in  air  are non-
      routine.   The techniques  developed  for  use in evaluating this procedure
      employ a 3-stage dynamic  gas dilution  system coupled with a constant-rate
      vapor generation  assembly containing  a trioxane permeation  tube   (VICI
      Medtronics Dynacal permeation  device or  equivalent) that  is maintained at
      55°C (See Appendix A).  Trioxane  vapor is converted stoichiometrically to
                                   8520 - 8                         Revision 0
                                                                  January 1995

-------
      formaldehyde  vapor using  a special
      converter assembly. (Ref.  4)
                       high-temperature  (160°C)  catalytic
      7.2   Determination of Volume to be Sampled - In this method, the volume
of ambient air to be sampled is dependent on the fact that the  air must  be mixed
at a certain rate with water.  This aqueous  solution is then mixed at a certain
rate with a reactant to form the chromophore which  gives an immediate  (after a
short time  delay)  and continuous  determination  of  the target  analyte.    The
sampling rate  set  for  this  equipment  has  been established at about 1.0 L/min.
See Sec. 8.4 and Table 3 for information regarding  tests to determine the bias
of the monitoring  system.
      7,3  Calculations

            7.3.1   Determination  of Equivalent Concentration,
      jug/m3, of Liquid Formaldehyde Standard Solution.
                                               in units of
                           (Cstd)(LFR)

                           (AFR)(AE)
            Where:
             "std
              LFR
              AFR
               AE
Concentration of standard solution in
Liquid flow rate calculated in Sec. 7.1.2 in mL/min
Air flow rate calculated in Sec. 7.1.1 in m3/nnn
Absorption efficiency of the air scrubber coil
(AE = 100% or 1.0 for this procedure)
            7.3.2   Determination  of  Formaldehyde  Concentration,  X,   in  Air
            Samples.
                     X  =

                  Where:

                     Y  .
                     b

                     m
                          (Y - b)
         m
    = Height of peak or plateau from analysis data.
    = The Y intercept of the calibration line from linear
      regression.
    = The slope of the calibration line from linear
      regression.
8.0  QUALITY CONTROL

      8.1  Refer  to  Chapter One and Method 8000  for  specific  quality control
procedures.
                                   8520 - 9
                                                Revision 0
                                              January 1995

-------
      8.2  Method  Blank - The analysis system is  zeroed using a regulated source
of zero grade air.  This "zero  air  supply"  allows  the analyzer  to establish a
baseline response, help zero the colorimeter, and calibrate the  system.

      8.3  Calibration Blank - Analyze a blank between each calibration standard.

      8.4  Precision and Bias

            8.4.1    The  precision  of  replicate  liquid  equivalent  standard
      analyses should  be ±5 percent, for the determination made within a 24 hour
      period.

            8.4.2   For day to day standard determinations,  the precision should
      be  better  than  ±10  percent  of  the  calculated  known   value  for  the
      standards. (Ref. 4)

            8.4.3    If  the analyzer performance  does not exceed  or  meet these
      requirements, instrument  recalibration procedures  are required.


9.0   METHOD PERFORMANCE

      9.1  Method  Sensitivity

            9.1.1   Accurate  determination of the sensitivity of  this procedure
      is complicated because of the difficulty involved  in  reproducing dynamic
      test  mixtures  at  less  than  12  /xg/m .   (Ref.   4)    Very  low  ambient
      concentrations,  on the order of 6 jug/m3,   produce a  measurable  response
      (about 10% of the full scale) on the analyzer;  however,  the precision of
      such measurements has  not been determined.

            9.1.2    As  the  volume  of  the  air  sampled  is  increased,  the
      sensitivity   of   the   method   increases  proportionally.    The  standard
      operating airflow rate   for  the  analyzer  is  between  0.5  L/min  and
      1.0 L/min.

            9.1.3    This  method  has demonstrated excellent  sensitivity  in
      applications  involving   the   monitoring   of  domestic    ambient   air
      environments. Data in  Table 1 are  from a test involving low-concentration
      formaldehyde measurements taken before and after changes  in the  heating,
      ventilation, and air conditioning  (HVAC) system  operation.

      9.2  Typical data obtained from a  multipoint calibration  of the  monitor,
using a 1.0 volt full-scale  output  range on  the  recorder and an airflow rate of
1.0 L/min on the analyzer,  is shown in Table 2.  (Ref.  4)

      9.3  Determination of Bias

            9.3.1   Tests performed  to measure the  bias of  this  procedure were
      conducted with  a laboratory  gas  dilution system  capable  of  generating
      dynamic test mixtures  of  formaldehyde  vapor  in air,  at flow rates up to
      5.0 L/min and concentrations in the range  from 12  to  500  jug/m3.  (Ref. 4)
      The  output  of  the  heated  permeation  tube  was determined,  at  weekly


                                   8520  -  10                        Revision 0
                                                                  January 1995

-------
      intervals, by accurate weighing on a calibrated analytical balance.  See
      Table 3 for test results.

            9.3.2  The results of the  tests  listed  in  Table 3 determined that
      the  accuracy  of the  permeation  tube  output  was on  the order  of ±3.5
      percent and the accuracy of the several flow calibration devices used to
      generate the test mixtures  is  on  the  order of ±5 percent.  These accuracy
      results do not indicate that the  modified CEA analyzer procedure displays
      any significant bias.


10.0  REFERENCES

1.    Miksch, R.R., Anthon,  D.W., Fanning,  L.Z.,  Hallowell,  C.D.,  Revzan,  K.,
      and Glanville, J., "Modified Pararosaniline Method for the Determination
      of Formaldehyde in Air," Analytical  Chemistry, Vol.  53,  1981,  pp.  2118-
      2123.

2.    Lyles,  G.R.,   Bowling,  F.B.,   and  Blanchard,  V.J.,   "Quantitative
      Determination  of  Formaldehyde   in   the  Parts  Per   Hundred  Million
      Concentration Level,"  Journal  of the Air Pollution  Control  Association,
      Vol.  15, 1965, pp. 106-108.

3.    West,   P.M.   and  Gaeke,   G.C.,   "Fixation   of   Sulfur  Dioxide   as
      Disulfitomercurate   (II)   and   Subsequent   Colorimetric   Estimation,"
      Analytical Chemistry, Vol.  28, 1956,  pp.  1816-1819.

4.    Fortune,  C.R.,  Daughtrey,  E.H.,   and  McClenny,  M.A.,  "Development of  a
      Portable Continuous Monitor for trace levels of Formaldehyde in Air," in
      Proceedings  of the  Annual Meeting   of  the  Air and Waste  Management
      Association, June 25-30, 1989, Anaheim, CA,  Paper No. 89.81.2.

5.    Instrument Operational  Manual, CEA Instruments,  Inc.,  16 Chestnut St.,
      P.O.Box 303, Emerson, N.J.  07630, Model TGM 555-FO.

6.    "Standard Test Method for Continuous  Measurement of Formaldehyde in Air,"
      Annual Book of ASTM Standards, D5221.
                                   8520  -  11                         Revision 0
                                                                  January 1995

-------
                                  TABLE  1.

         EXAMPLE  OF  AMBIENT  HCHO  MEASUREMENTS  OF  A  PRIVATE  RESIDENCE8
Analyzer Location
Kitchen/dining area
Kitchen/dining area
Kitchen/dining area
Bedroom area
Bedroom area
Time of Day
14:10
14:46
15:02
22:34
23:34
HCHO (fj.g/m3)
42.9
52.0
33.9
56.9
29.8
Conditions
HVAC on/house closed
HVAC off/house closed
HVAC off/house open
HVAC off/house closed
HVAC off/house open
     a  Measurements performed on 2-15-89.
                                  TABLE 2.

             MODIFIED  CEA MONITOR MULTIPOINT CALIBRATION DATA"-"
(For formaldehyde liquid standard concentrations versus CEA monitor readings)
HCHO Liquid Std.
Equivalent (^g/m3)
30.9
61.7
92.5
123.4
CEA Monitor Reading
(% of full scale)
26.1
51.7
77.2
101.3
Recorder Response
(mm)
65
129
193
253
  "  Analysis  performed  using  a  1.0 volt  full-scale  output  range  on  the
     recorder and an  air  flow rate of 1.0 L/min.
  b  Calculated regression coefficients  are slope = 0.8144, y-intercept =  1.26,
     and correlation  coefficient  (r) = 0.99989.
                                  8520 - 12
  Revision 0
January 1995

-------
                                TABLE 3.

          EXAMPLE OF CEA MONITOR DILUTION SYSTEM MEASUREMENTS8
Dilution
Primary
2
2
3
3
3
3
.000
.775
.800
.800
.800
.800
System Flow Rates (L/min)
Secondary



1
2
3
0
0
0
.000
.000
.000
Gas Removal
0
0
0
1.000
2.000
3.000
Total
2
2
4
4
4
4
.200
.975
.000
.000
.000
.000
HCHO
Calculated
124
92
68
51
34
17
.7
.3
.6
.4
.3
.2
Cone.
(uq/m3)
Measured %
124
87
65
48
30
10
.5
.9
.1
.2
.0
.9

Recovery
99.8
95.2
94.9
93.8
87.5
63.4
8  Measurements were performed on 2-7-89 with  the  standard delivery system
   operating at an output rate of 274 ±10 ng/min formaldehyde.
b  Regression  analysis   of  these  measurement   data   taken  at   6   air
   concentrations resulted in:  slope = 1.042 and correlation coefficient (r)
   = 0.99962.  The data are presented graphically in Figure 2.
                               8520 - 13
  Revision 0
January 1995

-------
                                                FIGURE  1.

                             FLOW  DIAGRAM  OF  FORMALDEHYDE  ANALYZER
(A) SAMPLE CELL     © AIR SCRUBBER

(£) REFERENCE CELL   © LIQUID/AIR SEPARATOR

© COLORIMETER    (£) GLASS DOUBLE MIXING TEE

    DRAIN TEE       (7) SINGLE 2S-TURN MIXING COIL


(tj GLASS DEBUBBLER (7) DOUBLE 11-TURN MIXING COIL

(7) PLASTIC TEE
DOUBLE « TURN TIME DELAY COIL

SINGLE 42-TURN TIME DELAY CON.


                   -fr-
                                                 8520  -  14
                                                 Revision 0
                                              January 1995

-------
                        FIGURE 2.

     CEA  MONITOR DILUTION MANIFOLD MEASUREMENTS
   105
   90
_:  75
o
3
5
60
   45
    IS
       LINEAR REGRESSION DATA (y • m« » b)
       N.6
       b. 5 180
       mm 1 042
       r•0 999S2
                         I
                               I
                                            I
           IS      30     45     60     75     90
                 HCHO CONCENTNATION (CALCULATCO). ppbv
                                               105
                        8520  - 15
                                                            Revision 0
                                                         January  1995

-------
                          METHOD 8520

CONTINUOUS MEASUREMENT OF  FORMALDEHYDE  IN  AMBIENT AIR
    7.1.1.1 Adjust air
    pump to produce a
  sample airflow rate of
      approximately
        1.0 L/min.
    7.1.1.2 - 7.1.1.3
    Calibrate airpump
      to the nearest
 1 .0 mL/min.  (The pump
  should  be re-calibrated
       as needed.)
   7.1.2 Calculate the
  actual liquid flow rate
  of the formaldehyde
   standard peristaltic
      pump tubing.
      7.1 .3 Prepare
   calibration standards
  and perform calibration
    of the  colonmter.
 7.2 Determine volume
    at ambient air to
      be sampled.
 Perform air sampling
as specified in Section
6.2, Routine Operating
      Procedure.
  7.3.1  Calculate the
  equivalent concen-
 tration  (CEQ ) of liquid
  from formaldehyde
  standard solution.
  7.3.2 Calculate the
  formaldehyde con-
   centration in the
     air samples.
                           8520  -  16
                          Revision 0
                       January  1995

-------
                                  APPENDIX A

                     FORMALDEHYDE  VAPOR  STANDARD GENERATOR

      A.I  A laboratory apparatus may be  constructed  to produce a continuous,
steady-state mixture of formaldehyde vapor at part-per-billion-by-volume (ppbv)
concentrations.  The apparatus consists  of a clean air generator system, a gas
dilution system,  and a  formaldehyde vapor generator.  These three components are
shown in Figures 1A, 2A, and 3A, respectively.

      A.2  Primary Dilution Air Supply

            A.2.1   An  in-house compressed air supply is connected to  a  dual
      silica-gel  cartridge assembly for moisture  removal (Figure  1A).   Use of
      the appropriate  valves  and  tube fittings permits  easy  replenishment of
      spent cartridges while maintaining continuous  operation of the system.

            A.2.2  The flow of dry air to the dilution manifold is regulated by
      a mass flow controller (Tylan Model  FC 260)  calibrated for a 0-5.0 L/min
      operating range.

            A.2.3  The dry air is  purified with a  400  cm3 capacity gas purifier
      (Alltech  Assoc.). The front half of the purifier is filled with activated
      charcoal  to remove organic  vapors, and the  back half  is filled  with 13X
      molecular sieve to remove any remaining  moisture and  formaldehyde vapor.

            A.2.4  The primary dilution  air from  this  system enters the three-
      stage gas dilution manifold (Figure 2A)  at  point (A).

      A.3  Three-Stage Dynamic Gas Dilution Manifold

            A.3.1  Construction of Manifold

                  A.3.1.1   The manifold  may be  constructed  by connecting three
            sections made  from high-purity quartz,  each of 10 mm  ID and about 50
            cm  length.   All connection ports are 1/4  in.  OD to  permit the use of
            standard  compression  fittings  (PTFE  Teflon)   to  assemble   the
            components.

                  A.3.1.2   All tubing  used throughout  the system is FEP Teflon
            (Cole-Parmer).  The 1/4 in.  union fitting  used to connect the first
            and second stages  of the manifold may  be modified by snugly fitting
            a short  length of  2  mm  ID Teflon  tubing  into  the union.   This
            restriction decreases  the size of the air passage by about one-half,
            which increases the air  velocity  significantly  at this point  to
            minimize any potential back-flushing of  the secondary dilution  air.
            The manifold  assembly was wrapped over its entire length  with  a
            resistive wire connected to  a  variable transformer (output  setting
            = 10%).

                  A.3.1.3   A K-type thermocouple may be attached to the  surface
            of  the  third  stage of the  manifold,  and the  entire assembly  is
            wrapped  with   a  1/2  in.  thick  foam insulation  material.    The


                                  8520  - 17                         Revision 0
                                                                  January  1995

-------
      temperature of the manifold should  remain constant at 30°C, varying
      by no more than ±1"C with changes in the room temperature.

            A.3.1.4  The exit end of the third-stage distribution manifold
      is  connected  to   a  4.0  L/min   rotameter  to  permit  easy  visual
      determination of the approximate  exit flow velocity.  The excess gas
      flow  is  continuously vented to  the fume hood  through  the exhaust
      manifold.

      A.3.2  Gas Flow Rates -  Gas flow  rates into and out of the manifold
are regulated  by  four calibrated mass flow controllers  [Tylan  Model  FC
260)  selected in  various flow  ranges up to  5.0 L-min   to  suit  each
requirement.  Flow rates are selected and monitored by using a Tylan Model
RO-28 readout/control unit.  Four 120 cm3 capacity gas purifiers (Alltech
Assoc.)  and four on/off valves  (Swagelock,  316  ss) are positioned  on
either side of the flow  controllers.

      A.3.3  Formaldehyde Dilution Scheme - The following serial dilution
scheme  is   versatile  and  responds  quickly  to adjustments  of  the  flow
systems;  that  is,   equilibration  is  achieved  within  1-3  min  after
adjustments  are  made.     More   importantly,  this  design  permits  the
generation  of  a tenfold-more-dilute dynamic mixture  than  a  single-stage
method using the same volume of gas.

            A.3.3.1  A continuous  stream of formaldehyde in N2  (200 cm3/cm)
      enters  the  first   stage of  the  manifold  at  point  (B)(Figure  2A),
      where it is mixed  with the  primary dilution air (up to 5.0 L/min).

            A.3.3.2   Near the  end of  the first stage, up to  90% of the
      primary gas mixture may be removed  by adjusting the flow controller
      to the  desired second stage  of  the dilution  manifold,  secondary
      dilution  of  the gas mixture  with  zero air  (< 5.0 L/min)  may  be
      performed in an analogous manner.

            A.3.3.3  The third stage of the dilution manifold is designed
      with  multiple  sampling  ports,  where the fully diluted  sample gas
      stream  (generally  flowing  at  3.0-5.0 L/min)  is available for test
      purposes.    The  second   mass  flow  controller/purifier  assembly
      connected to  the  house vacuum line  is  available for use  with  an
      optional extractive sampling method at flow rates  less than or equal
      to 2.0 L/min.

      A.3.4  The dilution manifold is connected to the formaldehyde vapor
generator and gas dilution system shown in Figure 3A.

A.4  Low-Concentration  Formaldehyde Vapor Generator/Gas Dilution System

      A.4.1    The  formaldehyde  vapor  generator  (Figure  3A)  may  be
fabricated  mostly  from  available spare  components and  some  custom-made
glass, quartz, and machined Teflon parts.

      A.4.2   A solid aluminum heater   block (10.7  cm diameter  x 25.4 cm
height)  is  drilled  to  house  the permeation tube  holder.   An  adjacent


                             8520 -  18                         Revision 0
                                                            January 1995

-------
      thermowell  is drilled  to  house  a  K-type  thermocouple  and  a  ceramic
      insulated RDT sensor for the temperature controller.  The aluminum block
      is wrapped with a heavy-duty (silicone rubber-coated) heat tape,  several
      coils  of  1/8 in.  copper  tubing  (to  preheat the N2  carrier  gas),  and a
      blanket of foam insulation.   This  assembly is mounted inside an insulated
      metal  box  (29  cm  length x 30 cm  18  cm  width  x 30  cm height).   A small
      opening is cut in  the hinged lid to permit removal of the permeation tube
      holder without opening the box.

            A.4.3  The temperature  controller (RFL Industries)  is mounted to the
      wall inside the insulated compartment to decrease the variability in the
      set point temperature fluctuations.   The externally  adjustable controller
      is set  to  maintain  a set point  temperature of  55°C (as  measured by the
      heater block thermocouple).

      A.5    Permeation  Tube  and   Holder   -  The  permeation   tube  holder  was
constructed from a borosilicate glass  vial  (2.5 cm ID x 12.5 cm height) with a
threaded opening.   A threaded  closure  with a  Viton  0-ring seal  was  custom-
machined from  a  solid  piece of PTFE Teflon.   The cap has two openings,  each
fitted with  a 1/8 in.  tube  fitting.   Preheated N2 enters the holder, passes
through a 10.0 cm length of 1/8 in. OD  FEP tubing,  and  sweeps the trioxane vapor
from the permeating tubeholder  into the phosphoric  acid  converter.   A Dynacal
permeation  tube   (5.0  cm  active  length)   containing  trioxane  (mp  64°C)  was
purchased from Vici Metronics (part No.  100-050-3520).  The nominal permeation
rate for this device is  680  ng/min cm  (±15%) at 90°C.   Periodic  gravimetric
determinations of weight  loss over  a period of 36 days  with this system resulted
in a measured rate of about 56 ng/min  cm  at 55°C.

      A.6  Converter

            A.6.1    The   high-temperature  converter  for depolymerizing  the
      trioxane vapor may  be  constructed from an  8.0  cm   section of  1.0  cm ID
      quartz tubing tapered at each end  to  1/4 in OD for tubing connections.  A
      7 mm diameter indention 2 mm deep was made at the center of the tube for
      attaching a temperature sensor  and thermocouple wire.

            A.6.2   The  converter  catalyst  is  prepared by saturating  14-mesh
      silicon carbide boiling  chips  (GFS  Chemicals) with 85%  phosphoric  acid
      (Fisher  Sci.  Co.);   saturation  was   ensured  by  heating  for  1 hour  and
      soaking  overnight.   The  quartz  tube  is  packed  tightly with  catalyst
      material, and small  portions of  quartz wool  (Refrasil)  were used to plug
      each end.

            A.6.3  The  converter tube  is wrapped with a nickel-chromium heater
      wire, a layer of  high-temperature fiberglass tape,  and  a double layer of
      fiberglass insulation wrap.   The leads of  the heater wire were connected
      to a  variable transformer  (output  setting =  16%)   that  is  wired to  a
      temperature  controller  (Omega  Engineering,  Model  6102-J)    The  ceramic
      insulated J-type sensor from the  controller and an exposed-junction K-type
      thermocouple are attached  to  the converter tube at the same point by using
      a silicon sealant.    The temperature  controller  set  point is  adjusted  to
      maintain the catalyst temperature  at  160°C.
                                  8520  -  19                         Revision 0
                                                                  January  1995

-------
            A.6.4  The outlet end of  the  converter  assembly is connected by a
      short  length  of  1/8  in.  tubing to  the  first stage  of  the  dilution
      manifold; see Figure 2A, point (B).

      A.7  Operation - The formaldehyde vapor generator/gas dilution system is
maintained in continuous  operation.  The primary dilution  air is normally set at
3.8 L/min and the N2 carrier gas at 200 cm3/min for  a continuous total  flow of
4.0 L/min.  An Omega Engineering digital thermocouple meter with a 10-position
selector switch (Model DSS-650) is used to monitor the three elevated temperature
zones in  the system,  the CEA monitor analytical  module, and  the ambient air
temperature.  A matched set of five exposed-junction K-type thermocouple wires
with PFA Teflon insulation (Omega Engineering) are used.
                                   8520  -  20                         Revision 0
                                                                  January 1995

-------
                                    FIGURE  1A

                           PRIMARY DILUTION AIR SUPPLY
    MASS FLOW
  CONTROL PANEL
          fTTTTm
   o  o   o  o
    ACTIVATED
    CHARCOAL
         13X   -v
      MOLECULAR VI
        SIEVE
                         PURIFIER
 DILUTION
MAMFOLD
                         PRIMARY
                         DILUTION
                            AIR
                                               QUICK
                                             CONNEa
                                              FITTINGS
"\J
                                             SILICAGEL
                                              DRYERS
                      ON/OFF
                      VALVE
                                          THREE-WAY
                                            VALVE •—}
                                                   /
               PRESSURE
              REGULATOR
             COMPRESSED
              AIR SUPPLY
                                   8520 -  21
                          Revision 0
                        January 1995

-------
TC®-AMBIENT AIR
TC® - DILUTION MANIFOLD
                                       FIGURE  2A

                     THREE-STAGE DYNAMIC GAS DILUTION  MANIFOLD
                                     TO EXHAUST HOOD
             TAMETER /
           (0-4.0 L/min) ;
                      u   £"~ijTOM~~/~>'
                         TO OPTIONAL SAMPLING METHOD
                                                                         FROM
                                                                     FORMALDEHYDE
                                                                       GENERATOR
                                       8520 - 22
  Revision 0
January 1995

-------
                                FIGURE 3A

LOW-CONCENTRATION FORMALDEHYDE VAPOR GENERATOR/GAS  DILUTION  SYSTEM
                                                       BH TO DILUTION MANIFOLD
         (160 'Q

(OPTIONAL SAMPLING METHOD
                                   TEMPERATURE
                                     NTROLLER
                                     (55 X)
                                                                »SP


Hill
o o o o
                                                     MASS FLOW
                                                   CONTROL PANEL
                                                                  ZERO
  (HCHO)}
 PERMEATION
   TUBE
                                                                          HOUSE
                                                                         VACUUM

                                                                          OP
                                                                           ZERO
                                                                           AIR
         HEATER
          BLOCK
                        110 VAC
                                        (INSULATED
                                           BOX
                                                                  TC® - PERM. TUBE HEATER
                                                                  TC® - H3PO« CONVERTER
                               8520 - 23
                                                                           Revision 0
                                                                         January  1995

-------
4.4  IMMUNOASSAY METHODS

         The following methods are included in this section:

         Method 4000:      Immunoassay
         Method 4010A:     Screening for Pentachlorophenol by Immunoassay
         Method 4015:      Screening for  2,4-Dichlorophenoxyacetic Acid by
                           Immunoassay
         Method 4020:      Screening   for  Polychlorinated   Biphenyls  by
                           Immunoassay
         Method 4030:      Soil  Screening  for  Petroleum Hydrocarbons  by
                           Immunoassay
         Method 4035:      Soil   Screening    for    Polynuclear   Aromatic
                           Hydrocarbons (PAHs) by Immunoassay
         Method 4040:      Soil Screening for Toxaphene by Immunoassay
         Method 4041:      Soil Screening for Chlordane by Immunoassay
         Method 4042:      Soil Screening for DDT by Immunoassay
         Method 4050:      TNT Explosives in Water  and  Soils by  Immunoassay
         Method 4051:      Hexahydro-l,3,5-trinitro-l,3,5-triazine (RDX) in
                           Soil and Water by Immunoassay
                                 FOUR - 15                       Revision 3
                                                               January 1995

-------
                                  METHOD 4000

                                  IMMUNOASSAY
1.0  SCOPE AND APPLICATION

      1.1  Immunoassay  is  an analytical technique  useful  for the separation,
detection  and  quantitation  of both organic  and  inorganic  analytes in diverse
environmental and waste matrices.  Immunoassay methods are used to produce two
types of quantitative results: 1) range-finding or  screening results indicative
of compliance with an action  level, and 2) assay values.

      1.2   Commercially-avail able testing products present immunoassay protocols
that are rapid, simple and portable.   These  products can  be used effectively in
both laboratory and  field  settings,  and require  limited training.  These test
products substantially increase the number of data  points that can  be generated
within  a  given time  period, and  permit  an operator  to analyze  a  number of
samples simultaneously, within a relatively  short  period of time.  Results are
available immediately upon completion  of the  test, and can assist  in the on-site
management of personnel  and  equipment, as well as  the data management activities
of the laboratory.

            1.2.1  A list of approved  immunoassay testing products  is available
     from the USEPA Office  of Solid Waste Methods  Section.

      1.3   Section 11.0 provides a glossary of basic immunoassay  terms.

            1.3.1   The  glossary is not intended  to be  comprehensive,  but to
      provide  basic  definitions  that  will  assist  in  understanding  product
      inserts and publications relating to immunoassay technology.

            1.3.2  The performance of  test products will  vary from manufacturer
      to manufacturer.   The  performance  claims  and limitations  of  each test
      product will  be provided  in the  package insert.   The  package  insert of
      each test product  purchased should be read to  determine  if the performance
      is acceptable for a given application.


2.0  SUMMARY OF METHOD

      2.1  The immunoassay test products available will often vary in both format
and chemistry.  The characteristics of  a specific product are described in the
package insert provided by the manufacturer.  This summary is, therefore, general
in scope, and is  intended to provide  a  general  description  of the more common
elements of these methods.

     Immunoassay test products use an antibody molecule to detect and quantitate
a  substance  in  a test  sample.   These  testing  products combine  the  specific
binding characteristics of an antibody molecule with  a detection chemistry that
produces a detectable response used for interpretation.   In  general,  antibody
molecules specific for the method's intended target  are  provided at  a predefined
concentration.  A reporter  (i.e.,  signal  generating)  reagent,  composed of the


                                   4000 -  1                          Revision 0
                                                                  January 1995

-------
target compound  conjugated  to  a signal producing compound  or  molecule (e.g.,
enzymes,   chromophores,  fluorophores,  luminescent  compounds,  etc.),   is  also
provided.   The  concentration,  affinity,  and specificity  of  the  products's
antibody influences performance, as  does  the chemistry of the reporter reagent.

     The reporter reagent  and antibody  molecules of a given product are binding
partners, and complex in solution.  The addition of a positive sample containing
the target substance to  this solution results  in a competitive binding reaction
for the  antibody sites.   The  antibody  concentration,  and  therefore  binding
capacity, is limited to prevent the simultaneous binding of both  the reporter and
sample molecules.   The concentration of  reporter reagent  that  can  bind to the
antibody is inversely proportional to the concentration of  substance in the test
sample.   Immunoassay methods may  be heterogeneous  (i.e., requiring  a  wash or
separation  step), or homogeneous  (i.e.,  not  requiring a  separation  step). In
commonly available heterogeneous testing products,  the antibody is immobilized
to a solid support such as  a  disposable  test tube,  and the bound  reporter reagent
will be  retained  after  removing the unbound  contents of  the tube  by washing.
Therefore, a negative sample results in the retention of more reporter molecules
than  a  positive  sample.     The  analysis  of   a  standard containing   a  known
concentration results in the immobilization of a  proportional concentration of
reporter reagent.   A positive  sample  (i.e., containing  a  higher concentration
than the  standard) results in the immobilization of fewer reporter molecules than
the standard, and a  negative sample  (i.e.,  containing  less  that  the  standard)
will immobilize more.

     2.2   A chemistry of  the detection of the immobilized reporter is used for
interpretation of  results.   The reporter  molecule may  be  a conjugate of the
target molecule  and a directly detectable chromophore, fluorophore,  or other
specie, or conjugated to an  enzyme that will act upon a substrate to produce the
detectable response.  Immunoassay testing products have a quantitative basis, and
will produce a signal that  is dependant on  the concentration of analyte present
in the sample.   For environmental  immunoassay methods,  the  signal  produced is
exponentially related to  the  concentration   of the  compounds  present.   Many
immunoassay methods use  enzymes to develop chromogenic response, and are termed
enzyme immunoassays.  Assays that generate a  chromogenic response are analyzed
photometrically,  and  use the principles of Beer's Law (Absorbance = Extinction
Coefficient x Concentration x  Path  Length) to determine  the concentration of
analyte in a sample.

Immunoassay methods can  provide quantitative data when configured with  a series
of reference standards that  are  analyzed and used  to construct a standard curve.
The signal  generated from  the  analysis of a  test sample  is  used  to  determine
concentration by  interpolation from the  standard curve.  Alternatively, these
testing  products  can be configured  to determine if  a  sample is  positive or
negative relative to  a single standard.

Individual immunoassay testing  products are reviewed and accepted by the EPA-OSW
for the detection of sample analytes  in  specified matrices. A variety of testing
products, produced by several different developers, may be available for  the  same
compound(s) and  matrices.   Each of theses methods have  been formulated using
independently developed  reagents  that  may result  in  significantly  different
performance characteristics  and limitations.
                                   4000 - 2                         Revision 0
                                                                  January 1995

-------
The performance of  the  immunoassay  testing  products  ultimately relates to the
characteristics  of  the  antibody,  reporter molecule,  and  sample  processing
chemistry.  The dose-response characteristics of a method, the position of the
standard relative to the claimed action level, and the stated cross-reactivity
characteristics  of the  selected  test  product,  provide  relevant  information
regarding the performance and recognition  profile of the selected test product.

The precision,  and  ultimately the sensitivity of an  immunoassay  method,  is a
function of the signal-to-noise  characteristics of its dose-response curve, and
its operational consistency.  Methods having a high slope and low non-specific
signal  generation  produce  the  most  sensitive  and precise  methods.   Signal
imprecision applied  to  a dose-response curve having  a  shallow slope exhibits
proportionally greater imprecision in the  calculated concentration than would a
method having a steeper slope.  In an action level  testing product,  this would
cause the  reference  standard to be positioned further  from  the  action level,
increasing the incidence of false  positive results.  Similarly, a method having
less non-specific signal  generation  (higher  signal-to-noise ratio) will be more
sensitive and precise when other characteristics  (/.e., dose-response slope) are
held constant.

Immunoassay methods are  used  to detect contamination at a specific concentration
below the  claimed  detection level   for the  test  product.   For example,  an
immunoassay used to  detect  PCB  contamination in soil  at  1 ppm will  include a
reference  preparation containing  less than  1 ppm.  The reference preparation
concentration is positioned to minimize  the  incidence of false negative results
at the claimed  detection level.   For remediation  and  monitoring applications,
where action levels of  interest  are  defined,  immunoassay methods should exhibit
a negligible incidence of false negative results,  and minimal false  positives.

For a single point action level  test, the concentration of analyte relative to
the action level is selected by the developer, and is influenced  by the precision
(i.e.,  intra-assay,  inter-person,  inter-lot, inter-day,  etc.),  sample matrix
interferences and other  performance characteristics and limitations of the basic
method.   The concentration  of analyte  in the  reference materials should be less
than,  but close  to, the claimed action  level.  The concentration  selected for the
standard defines the concentration  that will  produce  a  50% incidence of false
positive results by  the test product.   While this  issue  is  one  representing
limited liability to the operator, it is a practical  issue that often requires
attention.  An  immunoassay  method for  the  detection  of 1 ppm of PCB  using a
standard containing 0.8 ppm of PCB  will experience a 50% false positive incidence
in samples  containing  0.8 ppm  of PCB,  and  some  incidence of  false  positive
results in a sample containing between 0.8  and  1 ppm.  A similar immunoassay that
uses a standard  containing 0.4 ppm  will experience a 50% false positive incidence
in samples  containing  0.4 ppm  of PCB,  and  some  incidence of  false  positive
results in a sample containing between 0.4 and 1  ppm.   The closer the standard
concentration is to the action level, the better the  overall  performance.

      2.3  Cross reactivity  characteristics  illustrate  the specificity of the
underlying immunochemistry.   The antibody  molecules used by a test product bind
to a target compound and then  participate in the process of  generating the signal
used  for   interpretation.      Antibody  molecules   bind  by   conformational
complimentarity.    These  molecules  can  be  exquisitely  specific,   and  can
differentiate subtle differences  in the structure of  a  compound.  The  binding


                                   4000 -  3                          Revision 0
                                                                  January 1995

-------
characteristics of reagents in different test products can vary, and influence
the recognition profile and incidence of false results obtained by the method.
Immunoassay methods should detect the target analytes claimed by the test product
and exhibit limited recognition for compounds and substances not specified.


3.0  INTERFERENCES

      3.1  Non-target analytes may bind with the antibody present,  producing a
false-positive result.  These non-target analytes may be similar to the target
analytes,  or  they may  be  chemically  dissimilar   co-contaminants.    During
evaluation  of  each test product  for  RCRA testing applications,  studies  were
conducted to determine these "cross-reactive" constituents.  At a minimum, these
studies evaluated  the  response  of the test product  to  all  other  similar  RCRA
analytes  in  that  analyte  class,  as  well  as for  selected lists  of  non-RCRA
analytes.  This testing scheme is designed to ensure that all other similar RCRA
analytes  and  likely  co-contaminants  are  evaluated during  cross-reactivity
testing.  The results of these studies are presented  in each method in tabular
form, providing separate data sets for each test product evaluated.

      3.2   Interference in the binding of an antibody to its target compound,
or  reporter molecule  reagent,  may  occur  when  testing sample matrices  with
confounding  contaminants  or circumstances  (e.g.,  oil,  pH,  temperature,  some
solvents).  Immunoassay products contain sample processing technology that has
been developed  and validated  for use with  specified  matrices.   Interferences
incurred  from  the testing  of  incompatible matrices  may prevent  the  testing
product from meeting  its performance  claims,  and increase the number  of false
positive or false  negative results.   Individual  immunoassay products designate
the intended sample matrices.

      3.3   Immunoassay products differ in shelf-life  and storage requirements.
Test products that  are operated outside of the shelf-life  and  storage temperature
recommendations may not provide the claimed performance.

      3.4   Some test products have designated temperature ranges for operation.
When these products are used,  all tests must be performed within the specified
operating temperature limits, or else false negative/positive results may exceed
performance claims.


4.0  APPARATUS AND MATERIALS

      4.1  Each test product will  specify  the  apparatus  and materials provided,
as well as any additional apparatus and materials necessary for performance of
the test.
5.0  REAGENTS

      5.1  The two basic reagents used in  immunoassay  analysis are the antibody
(e.g., anti-PCP) and reporter conjugate reagent (e.g.,  PCP molecules  bound to an
enzyme).
                                   4000 - 4                         Revision 0
                                                                  January 1995

-------
            5.1.1 The formation of antibodies to  haptenic molecules  (i.e., most
      environmental contaminants) is induced by the derivatization and coupling
      of molecules  of  the target analytes  to large  carrier  molecules such as
      albumin, hemocyanin or thyroglobulin.  The increased size and complexity
      of the  immunogen (antigen) conjugate,  once  injected,   is  sufficient to
      stimulate  the  immune  system  to  produce an  antibody  response.  The
      effectiveness  of  the  immunogen  in  producing   antibodies   having  the
      prerequisite binding characteristics and recognition profile  is influenced
      by the surface density of the chemical groups  on the carrier molecule, the
      nature  of  the  bridge  chemistry  used,  the  point  of attachment,  the
      immunization  protocol,  •immunogen  concentration,  adjuvants  (i.e.,  immune
      response stimulants), and the species of the host animal.

            5.1.2 An  enzyme-reporter  conjugate  reagent  is  synthesized  by
      coupling a target analyte or derivative of  a target analyte to an enzyme,
      such as horseradish  peroxidase.   Enzymes  enhance  the sensitivity of the
      method  by  action  on  a  substrate  and  the   production  and  catalytic
      amplification of the  detection signal.  A  single enzyme molecule used in
      immunoassay methods will convert approximately 106 molecules of a target
      analyte  into  a  detectable  product  within   one   minute   at  ambient
      temperature.

      5.2  Each test product will  specify the reagents provided, as  well as any
additional  reagents necessary for performance of the test.


6.0  SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1  Testing  of  solid  waste  by  immunoassay   requires  production  of  a
reproducible, particulate free leachate.  It is  critical that this leachate be
produced using a  solvent that allows the reproducible  extraction and  recovery of
the target  analytes, and is  compatible with the antibody/enzyme conjugate of the
immunoassay system used.   Buffers, detergents, and solvents, used together  or in
combination,   have  been  used  effectively  for  extraction.    Filtration  of
particulate matter may  be  integrated into  the  immunoassay test, or accomplished
as a separate step within the protocol.

      6.2  The immunoassay test products included in  SW-846  methods will provide
explicit waste- or medium-specific directions for  handling samples and extraction
of target analytes.

      6.3  See the  introductory material  to this  chapter,  Organic Analytes,
Section 4.1.
7.0  PROCEDURE

     7.1   The specific procedure for each  immunoassay test product is supplied
by the manufacturer in the package insert.

      7.2   The recognition characteristics, sensitivity, detection ranges(s),
effective operating temperature,  interferences  and cross-reactivity of the test
will  depend on the product being used.


                                   4000 - 5                         Revision 0
                                                                  January 1995

-------
      7.3   Immunoassay methods include both a sample processing and immunoassay
component.   It should  be  noted that the  immunochemical  reagents  and  sample
processing  components  supplied  with   each   product   is   specific  to  each
manufacturer.   Methods available  from   different  manufacturers  for  the  same
compound  and   application   may  have   significantly   different   performance
characteristics.
8.0  QUALITY CONTROL

      8.1   The performance of the tests cited in the immunoassay methods  in this
manual has been reviewed,  and found  to  be  consistent  with  the claims that are
made  in  the manufacturer's  literature.   In order  to  meet  this  performance
expectation, the analyst must:

      o    Follow the  manufacturer's  instructions  for the  test  product being
used,
      o    Use test products before the specified expiration date,
      o    Use  reagents  only  with  the  test  products  for   which  they  are
           designated,
      o    Use the test products  within  their specified storage temperature and
           operating temperature limits.

      8.2   It is important to evaluate the  performance  claims and limitation
provided with each  testing  product  to  determine  its  application to a specific
matrix and testing program.

     8.3    Refer to Chapter One for standard quality control  procedures.


9.0  METHOD PERFORMANCE

      9.1   A false  negative  is  defined as  a negative response for  a  sample
containing  the  target  analytes  at or  above the  stated  action level.  False
negative rate is measured  by analyzing split samples using both the test product
and a separate reference method.   False negative data are provided in each method
for each test product evaluated.

      9.2   A false positive is defined  as a positive response  for a sample that
contains analytes  below the  specified action level.   Like false  negatives, false
positive rates  are measured by analyzing  split samples with both the test product
and a separate reference method.   False positive data are provided in each method
for each test product evaluated.

      9.3   Cross-reactivity and  recognition profile data are provided at  the end
of  each  method  in  tabular  form, providing  separate  data  sets for  each test
product evaluated.  Using these data, the analyst can evaluate if contaminants
are  present  which  are  likely  to produce  a  false  negative response,  and the
magnitude of that response.

      9.4   For single-point tests,  sensitivity data are provided demonstrating
the concentration  of target analyte(s) that  can be detected with greater than 95%
confidence.
                                   4000 - 6                         Revision 0
                                                                  January 1995

-------
      9.5   Data  are  provided  demonstrating the bias  of  the testing products
accepted.  These data may be from:

     o     serial dilution of samples (i.e., is the recovery of target analyte
           a function of concentration?),
     o     sample recovery studies, and
     o     studies  correlating  the  results  of the  testing  product with  a
           reference method.

      9.6  Data are provided demonstrating that the extraction efficiency  of the
test being evaluated correlates with that of the referenced method.


10.0 REFERENCES

1.   S.B. Friedman, "Doing Immunoassays  in  the  Field",  Chemtech, December 1992,
     pp 732-737.

2.   Roitt, L.,  Brosstoff,  J.,  Male, M.,  (eds.),  Immunology,  J.B. Lippincott
     Co., Philadelphia, Pennsylvania, 1989

3.   Stites, Daniel P., Terr,  Abba  I.,  (eds.),  Basic and Clinical Immunology,
     Appleton and Lange, Norwalk, Conneticut, 1991

4.   Odell, W.D. and Daughaday,  W.H., Principles of Competitive Protein-Binding
     Assays, J.B. Lippincott Co., Philadelphia, Pennsylvania, 1971

5.   Ishikawa, E., Kawai, T., Miyai, K.  (eds.),  Enzyme  Immunoassay,  Igaku-Shoin,
     Tokyo, Japan, 1981

6.   Tijssen,  P.  (ed.), Practice and Theory of Enzyme  Immunoassays, Volume 15,
     Elsevier, NY, NY, 1985

7.   Butler,  John  E.   (ed.),  Immunochemistry  of Solid-Phase  Immunoassay,  CRC
     Press, Boca Raton, Florida, 1991

8.   Ngo, T.T.,  Lenhoff, H.M.,  Enzyme-Mediated  Immunoassay,  Plenum Press,  New
     York, 1985

9.   510K of the Federal Food,  Drug and Cosmetics Act, Section 21, CFR 807.87


11.0 GLOSSARY OF TERMS

Antigen                 A  molecule   that  induces  the  formation   of  an
                        antibody.

Antibody                A binding protein which  is  produced  in response
                        to an antigen,  and which has the ability to bond
                        with the antigen that stimulated its production.
                                   4000 - 7                         Revision 0
                                                                  January 1995

-------
B Lymphocyte
(B Cell)
Carrier
Competitive
Immunoassay

Cross-Reactivity
Dose-Response
Curve
ELISA
Enzyme Conjugate
Enzyme Immunoassay
A  type of  lymphocyte
differentiates into an
cell.
 that,  upon  stimulation,
antibody-secreting plasma
An immunogenic substance that,  when coupled to a
hapten, renders the hapten immunogenic.
An  immunoassay  method  involving  an
competitive binding reaction.
                 in-vitro
The  relative  concentration  of  an  untargeted
substance   that  would   produce   a   response
equivalent to  a specified concentration of  the
targeted  compound.      In   a   semi-quantitative
immunoassay,   it  provides an  indication of  the
concentration   of   cross-reactant  that   would
produce  a  positive  response.    Cross-reactivity
for individual  compounds  is often  calculated  as
the ratio  of target substance  concentration  to
the cross-reacting substance concentration at 50%
inhibition of the immunoassay's maximum signal  X
100%.

Representation  of  the  signal   generated  by  an
immunoassay   (y  axis)   plotted    against   the
concentration of the target  compound (x axis)  in
a  series of  standards  of  known  concentration.
When  plotting  a  competitive   immunoassay  in  a
rectilinear format,  the dose-response will have a
hyperbolic   character.     When   the   Iog10   of
concentration   is  used,  the   plot  assumes   a
sigmoidal shape, and  when  the  log  of  signal  is
plotted  against  the  logit  transformation   of
concentration, a straight line  plot is produced.

Enzyme  Linked  Immunosorbent Assay  is  an  enzyme
immunoassay  method   that  uses  an  immobilized
reagent  (e.g.,antibody   adsorbed  to  a  plastic
tube), to facilitate the  separation of targeted
analytes  (antibody-bound
target  substances   (free
using a washing step, and
generate the  signal   used
of results.
                                                  components)  from  non-
                                                   reaction  components)
                                                  an enzyme conjugate to
                                                  for  the  interpretation
A molecule produced by the coupling of an enzyme
molecule  to  an  immunoassay  component  that  is
responsible  for acting    upon  a  substrate  to
produce a detectable signal.

An  immunoassay  method   that   uses   an   enzyme
conjugate reagent to generate  the signal used for
interpretation of results.  The  enzyme  mediated
                                   4000 -  8
                                            Revision 0
                                          January 1995

-------
False Negatives
False Positives
Hapten
Hapten-Carrier
Conjugate
Heterogeneous
Immunoassay
Methods

Homogeneous
Immunoassay
Methods
Immunoassay
Immunogen
Ligand
response  may  take  the  form  of a  chromogenic,
fluorogenic, chemiluminescent  or  potentiometric
reaction, (see Immunoassay and ELISA)

A   negative   interpretation    of   the   method
containing  the  target  analytes at or  above  the
detection level.   Ideally,  an  immunoassay  test
product  included  in  an  SW-846  method  should
produce   no   false  negatives.    The   maximum
permissible  false  negative   rate  is   5%,   as
measured  by  analyzing  split samples using  both
the test product and a reference method.

A positive interpretation  for a  sample is defined
as a positive  response for a sample that contains
analytes below the action  level.

A substance that cannot directly induce an immune
response  (e.g.,  antibody production),  but  can
bind to the  products of an immune response (e.g.,
antibody) when  that  response  is  induced by  an
alternate mechanism.   Chemical contaminants  of
the environment are haptens.

The coupling of a non-immunogenic molecule (e.g.,
targeted  analyte)  to  an  immunogenic  substance
(e.g.,  bovine  serum  albumin,  keyhole  limpet
hemocyanin)  for the purpose  of stimulating  an
immune response.

Immunoassay methods  that  include  steps  for  the
separation of substances that become bound to the
antibody from  those that remain  free in  solution.

Immunoassay  methods  that   do   not  require  the
separation of  bound and free substances, but that
utilize  antibody   molecules  that  can  bind  and
directly  modulate  the  signal   produced  by  the
reporter molecule  (e.g.,  enzyme conjugate).

An  analytical  technique  that  uses  an  antibody
molecule as a binding agent  in  the  detection  and
quantitation  of substances  in a  sample,  (see
Enzyme Immunoassay and ELISA)

A substance  having  a minimum size and complexity,
and that is  sufficiently foreign to a genetically
competent host to stimulate  an  immune  response.
The molecule, ion or group
with another molecule.
that forms a complex
                                   4000  -  9
                                            Revision  0
                                          January  1995

-------
Lymphocytes             One  of  the  five  classes  of  white  blood  cells
                        found in the  circulatory  system  of vertebrates.
                        A mononuclear cell  7-12 pm  in diameter containing
                        a  nucleus  with densely  packed  chromatin  and  a
                        small rim of cytoplasm.

Monocl onal      Identical copies of antibody molecules that have
Antibodies              a common set of binding  characteristics.

Polyclonal      A  group of  antibody molecules  that differ  in
Antibodies              amino acid  composition   and  sequence,  and  that
                        exhibit  binding  characteristics.     Polyclonal
                        antibodies  are produced  from  a   simulation  of
                        multiple clones of lymphocytes.
                                   4000  -  10                        Revision 0
                                                                  January 1995

-------
                                 METHOD 4010A

                SCREENING FOR PENTACHLOROPHENOL BY IMMUNOASSAY
1.0  SCOPE AND APPLICATION

      1.1  Method  4010 is  a procedure  for screening  solids such  as soils,
sludges,   and   aqueous  media   such  as   waste   water  and   leachates   for
pentachlorophenol  (PCP) (CAS Registry 87-86-5).

      1.2  Method 4010 is recommended for screening samples to  determine whether
PCP is likely to be present at  defined concentrations  (i.e., kits are available
which give positive results at 0.005  mg/L for aqueous samples,  and at  0.5, 10 or
100  mg/kg  in  solid  samples).    Method  4010   provides  an  estimate  for  the
concentration of PCP by comparison with a standard.

      1.3  Using the  test  kits from which  this method was developed,  95% of
aqueous samples containing 2 ppb or  less of PCP will produce a negative result
in the 5 ppb configuration.  Also,  95% of soil samples containing 125 ppb or less
of PCP will produce a negative result in the 5000 ppb test configuration.

      1.4  In cases where the exact concentration of PCP is required, additional
techniques (i.e.,  gas  chromatography (Method 8040)  or gas chromatography/mass
spectrometry (Method 8270))  should be used.

      1.5  This method  is  restricted to  use  by  or under the  supervision  of
trained  analysts.    Each   analyst  must  demonstrate the  ability   to  generate
acceptable results with this method.


2.0  SUMMARY OF METHOD

     2.1   Test  kits  are  commercially  available  for  this  method.    The
manufacturer's directions should be followed.

     2.2   In general,  the method is performed using  a water sample or  an extract
of a water sample.  The sample/extract and an enzyme  conjugate  reagent are added
to immobilized  antibody.  The enzyme conjugate "competes" with PCP present in the
sample for binding  to immobilized anti-PCP antibody.  The test  is interpreted by
comparing the response produced by  testing  a sample  to the response produced by
testing standard(s) simultaneously.


3.0  INTERFERENCES

      3.1  Compounds that are chemically similar may  cause a positive test (false
positive)   for  PCP.  The  test  kits  used  in preparation  of this  method  were
evaluated  for  interferences.   Tables 1A  and  IB provide the  concentration  of
compounds which will  give a false positive test  at the indicated concentration.

      3.2  Other compounds  have been tested for cross reactivity for  PCP and have
been demonstrated not to interfere with  the specific kits  tested.   Consult  the
information  provided  by  the  manufacturer of  the  kit  used for  additional

                                    4010A-1                          Revision  1
                                                                  January  1995

-------
information regarding cross reactivity with other compounds.

     3.3   Storage and use temperatures may modify the method performance. Follow
the manufacturer's directions for storage and use.


4.0  APPARATUS AND MATERIALS

      4.1  Immunoassay test kit: PENTA RISc™  (EnSys, Inc.), EnviroGard™ PCP in
Soil (Millipore,  Inc.), or equivalent.   Each commercially available test kit will
supply or specify the apparatus and materials necessary for successful completion
of the test.
5.0  REAGENTS

     5.1   Each  commercially  available test  kit will  supply or  specify  the
reagents necessary for successful completion of the test.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1   See the  introductory material  to  this chapter,  Organic  Analytes,
Section 4.1.

     6.2   Soil samples may be contaminated, and  should therefore be considered
hazardous and handled accordingly.


7.0  PROCEDURE

     7.1   Follow the manufacturer's instructions for the test kit being used.
Those test kits used must meet  or exceed the performance specifications indicated
in Tables 2-10.
8.0  QUALITY CONTROL

     8.1   Follow the manufacturer's instructions for the test kit being used for
quality control  procedures specific to the test kit used. Additionally, guidance
provided in Method 4000 and Chapter One should be followed.

     8.2   Use  of  replicate  analyses,   particularly  when  results   indicate
concentrations  near  the action  level,  is  recommended to  refine  information
gathered with the kit.

     8.3   Do not use test kits past their expiration date.

     8.4   Do not use tubes  or reagents designated for use with  other test kits.

     8.5   Use  the  test kits  within  their specified  storage  temperature and
operating temperature limits.
                                    4010A-2                         Revision 1
                                                                  January 1995

-------
     8.6   Method 4010 is intended for field or laboratory use.  The  appropriate
level of quality  assurance  should accompany  the application of this method  to
document data quality.


9.0  METHOD PERFORMANCE

      9.1  This method  has been  applied  to  a  series  of groundwater,  process
water,  and wastewater samples from industries which  use PCP, and  the  results
compared with  GC/MS  determination  of  PCP (Method  8270).    These results are
provided in Table  2.   These results represent  determinations by two laboratories
using the PENTA RISc™ test kit.

      9.2  This method  has  been applied to a  series  of soils from  industries
which use PCP and the  results compared with GC/MS determination of PCP via method
8270.    These  results  are  provided  in  Table  3.    These  results represent
determinations by two laboratories using the PENTA RISc™ test kit.

      9.3  Sensitivity of the EnviroGard PCP in Soil Test Kit  was  determined  by
establishing the  "noise"  level  expected from  matrix  effects  encountered  in
negative soil  samples and determining the corresponding TPH concentration  by
comparison to the  analyte-specific response curve.   Eight different  soils which
did  not contain  PCP were  assayed.    Each  of  these  soils was  extracted  in
triplicate and  each  extract was analyzed in three different assays.  The  mean and
the standard deviation of the resulting %Bo's  (%Bo =  [(OD8ample/ODnegatlvecontrol)xlOO])
were calculated and the sensitivity was estimated  at two  standard  deviations
below the mean.  The sensitivity for Method 4010 was determined to be 80% Bo  at
a 95% confidence  interval.   Based on the average  assay response  to PCP,   this
corresponds to 2 ppm  PCP.  These data are shown in Table 4.

     9.4   The effect of water content of the soil samples on the EnviroGard™ PCP
in Soil  test kit was  determined by assaying three different  soil  samples which
had been dried  and subsequently  had water added to 30% (w/w).  Aliquots  of these
samples were then  fortified with PCP.  Each soil  sample was  assayed three times,
with  and without  added  water,  and  with  and without home  heating oil (HHO)
fortification.   It was determined  that water in soil up to 30% had  no detectable
effect on the method.  These data are shown in Table  5.

     9.5   The effect of the  pH of  the  soil  extract on the EnviroGard™ PCP  in
Soil test kit  was determined by adjusting the  soil pH of  three  soil  samples.
Soil samples were  adjusted to  pH 2 - 4 using 6N HC1 and pH 10 - 12 using  6N NaOH.
Aliquots of  the  pH  adjusted  soil  samples were  fortified  with  PCP  and  the
unfortified and fortified samples were extracted.  These extracts were  assayed
three times.  It was  determined that soil samples with pH ranging from  3 to  11
had no detectable  effect on the performance of the method.  These data are shown
in Table 6.

     9.6   The bias  of  the  EnviroGard™  PCP  in Soil test kit was  estimated  by
fortifying three different soil  samples  at two different  concentrations  (10 and
100 ppm  PCP).   Each  fortified sample  was extracted  three  times  and each  extract
was assayed  three  times.  Recovery for individual determinations ranged  from 60%
to 125%.  Average  recovery for each  individual extract ranged  from 72% to 101%.
Overall  average recovery for all  samples was 86%.  These  data  are  summarized  in


                                    4010A-3                           Revision 1
                                                                  January 1995

-------
Table 7.

     9.7   The  effect  of  co-contamination  of  soil  samples with  oil  on  the
EnviroGard™ PCP  in  Soil  test kit was  investigated.   Three  soil  samples  were
adulterated with diesel oil and aliquots were fortified with PCP.  The samples
were extracted and the  extracts each assayed three times.  It was determined that
no interference was detected in samples with up to 10% oil contamination.   The
data from samples adulterated at 10% are shown in Table 8.

     9.8   A  field  trial  was  conducted at  a  contaminated  site  using  the
EnviroGard™ PCP in Soil test kit.  Method 4010 was used to identify soil which
had been contaminated  with PCP  from wood  treatment operations.   A total of 33
samples were analyzed  including 5 field duplicates.   For the field duplicates,
the reference method demonstrated an average  coefficient of variation of 16%.
For Method 4010 average coefficient of  variation  was 31%.  Since Method 4010 is
not quantitative, quantitative values were estimated.   These data are shown in
Table 9.  At the 10 ppm  cutoff,  there  were  0/33  (0%)  false  negatives  and 0/33
(0%) false positives.   At the 100 ppm cutoff, there was  1/33 (3%) false negatives
and 1/33 (3%)  false positives.  These data are shown  in Table 10.


10.0 REFERENCES

1.   J.P. Mapes, K.D. McKenzie, L.R. McClelland, S. Movassaghi,  R.A. Reddy, R.L.
     Allen,   and  S.B.   Friedman,   "Rapid,   On-Site   Screening  Test   for
     Pentachlorophenol  in Soil and Water - PENTA-RISc™", Ensys Inc.,  Research
     Triangle Park,  NC 27709

2.   J.P. Mapes, K.D. McKenzie, L.R. McClelland, S. Movassaghi,  R.A. Reddy, R.L.
     Allen, and S.B.  Friedman,  "PENTA-RISc™  - An  On-Site  Immunoassay  for
     Pentachlorophenol  in Soil", Ensys  Inc., Research  Triangle Park,  NC 27709

3.   PENTA-RISc™ Instructions for Use, Ensys Inc.

4.   EnviroGard™ PCP  in Soil  Test Kit  Guide,  Millipore,  Inc.
                                    4010A-4                         Revision 1
                                                                  January 1995

-------
Table 1A - Cross Reactivity for POP
PENTA RISc™ Test Kit
Compound3
2,6-Dichlorophenol
2,3,4-Trichlorophenol
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
2,3,5,6-Tetrachlorophenol
Tetrachlorohydroquinone
Concentration (mg/Kg)
in Soil to Cause a
False Positive for
PCP at 0.5 mg/Kg
700
400
16
100
1.2
500
Concentration (/ng/L)
in Water to Cause a
False Positive for
PCP at 5 jiig/L
600
600
100
500
7
>1500
a Compounds assayed at 3.75 /LtM  (molar  equivalent  of PCP at 1000 M9/L)> except
where noted.
Table IB - Cross Reactivity for PCP
EnviroGard™ PCP in Soil Test Kit
Compound
Pentachlorophenol
2,5-Dichlorophenol
2,6-Dichlorophenol
2,3,4-Trichlorophenol
2,3,5-Trichlorophenol
2,3,6-Trichlorophenol
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
Lower Limit of
Detection (mg/kg)
10
1000
1000
1000
500
500
500
500
The following compounds were tested and found to yield
negative results at 1,000 ppm:
2,3,5,6-Tetrachloronitrobenzene PCB (Aroclor 1248)
3,5-Dichlorophenol TNT
2,4-Dichlorophenol DDT
2,3-Dichlorophenol PAHs
4-Chlorophenol Chlordane
                                    4010A-5
  Revision 1
January 1995

-------






X
•1 —
S-
+J
ro
SI
ueous
cr

i

00
\
5
3
£
00
O)
1 —
h-
00
C£

"^
LjJ
Q.
<4_
O
c
O
00
-r~
ro
CL
0
,
CNJ
QJ
,^
JD
ro
1 —





03 __
1— ^
^y Ll_
UJ
i . i U_
C3 ~
-
d)
5-
3
00 *-"—s
re E
(!) Q.
E CL
C
o oo
•-- s:
rO l_)
S- C3
C >,
d) fi
O
rr
0





Q.
Q.
- — '
00
•*->

Ol
o;
en
c:
'^
O)
0)
s_
o
oo







o

LO


1 — 1


in
•
o



^
o



LO
CD
O
O
O
•
O
O)
Q.
1—
O)
Q.
E
oo



a.
LL_









LO
CO






V

A


A




















i-
ai
ro
s
"^
o
S-
cn



>•









LO
oo
o















V




A


















>•









1— 1
o
V


























V


A









>•









00






V

A






























>•









00
CVJ








V


A














A


A









7-









OJ








V


A














A


A









g









*— i
C\J
o















V




V





A


A


a>
5
00
00
•









r*.
i — i
o















V




V





A


A









>-









CM
t — i
0















V




A


















D-
U_









O











V



V




A





A





i-
OJ
ro
3
CD
"^~*
ro
3



s-









i — i








V


A



























3-









i — I
O
V




















V





A


A









>-









i — i
O















V




A


















>-









i — i
O
V




















V





A


A


4_
M-
0
C
3
*"



5-









ro
0
O


























V


A









5-









00
0
0











V














A












>-









oo
0
0











V














A












>-









00
CD
0











V














A












CL.
u.









f-H
CVJ








A




























            €
•—i  IT)
    CTl
 C  CTl
 O  ^^

 00   >,
•r-   S_
 >   ro
 a>   =j
O
1—I
o

-------






X
y
i/>
3
O
Q)
3
CT
**•
I

CO
s:
5
i
2
00
QJ
1 —
2
u
CO
^
I—
LlJ
Q.
4-
O
c
O
.^
03
CL
O
1
CNJ
QJ
r_
jQ
ro
h-






ro z
LU
S ^
LU u-
C£
<-s ^
"
Q)
3
CD Q.
C
O OO
••- 2:
n3 <_5
S- CJ3
C >,
QJ ^
O
C
O





-g-
Q-
Q.
• —
i/l

Q)
Qi
CD
c
'^
QJ
QJ
CO







O
LO

LO



r-H


IT)
O



.
O

in
o
o
in
0
o
.
0
CD
Q.
[-__
p _
OJ
Q.
ro
CO

>-






ro
o
o












V











A













Q_
U_






Osl
o
o












V











A













>-






ID
0
O
O
























V



A









LL.






cn
ID
o












A

























>.






cn
CO
CO
o












V











A













Q.
U.






0
i — i
O












A

























Q_
U_






i — I
r-H
O












A

























^






CO
o








V











A

















>-






r-H









A





























>-






cn
i — i









A





























^






CO







V





A
























i— i un
    CTl
 c o>
 O r-1
 >  CO
 CD  3
ct:  c
O

o

-------
Table 3
Comparison of PENTA RISc™ Test Kit with GC/MS
Soil Matrix
Concentration measured
by GC/MS (ppm)
1100
88
0.31
0.72
315
1.5
6.4
9
1.9
46
<1
21
3.3
4
11
18
33
54
65
74
83
1.1
14.3
<1
<1
<1
Screening
Results (ppm)
0.5
>
>
<
<
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
<
<
<
5
>
>
<
<
>
<
>
>
<
>
<
>
<
>
>
>
>
>
>
>
>
<
>
<
<
<
50
>
<
<
<
>
<
<
<
<
>
<
<
<
<
<
<
<
>
>
>
>
<
<
<
<
<
AGREEMENT3
Y, FP, FN
Y
FN
Y
FN
Y
Y
Y
Y
Y
FP
Y
Y
Y
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
4010A-8
  Revision 1
January 1995

-------
Table 3
Comparison of PENTA RISc™ Test Kit with GC/MS
Soil Matrix
Concentration measured
by GC/MS (ppm)
3.9
<1
1.4
48
<1
142
108
117
56
2.5
3.5
143
nd
0.02
5
Screening
Results (ppm)
0.5
>
<
> j
>
<
>
>
>
>
>
>
>
<
<
>
5
<
<
<
>
<
>
>
>
>
<
>
>
<
<
<
50
<
<
<
>
<
>
<
>
>
<
<
>
<
<
<
AGREEMENT3
Y, FP, FN
Y
Y
Y
FP
Y
Y
FN
Y
Y
Y
FP
Y
Y
Y
Y
4010A-9
  Revision 1
January 1995

-------
                                    TABLE  4
                 EnviroGard™  PCP  in Soil Test  Kit3  Sensitivity

                 Part 1 - Average Response with  Negative Soils
Soil#     Soil  Type    Average %Bo (n = 9)Standard Deviation                          ^^
SI
S2
S3
S4
S5
S6
S7
S8
LOAM
CLAY
SAND
LOAM
SAND
CLAY
LOAM/SAND
SAND/LOAM
97.6
100.1
101.4
99.4
100.2
97.4
102.6
97.5
3.0
1.4
2.8
4.9
3.1
2.7
0.3
3.6
AVERAGE                   99.5               5.2


         Part 2  - Average Response with Pentachlorophenol Calibrators

PCP
Concentration (ppm)  Average AbsorbanceAverage %Bo
0
5
20
50
200
1.142
0.828
0.556
0.382
0.162
N/A
72.6
48.7
33.4
14.1
                          Part 3  - Method  Sensitivity
Based on Part  1  and  Part 2 Above:
         Average %Bo -  2 SO = 89.2 which is equivalent to  1.6  ppm  PCP
         Average %Bo -  3 SD = 84.0 which is equivalent to  2.3  ppm  PCP
(%Bo = IIODsample/ODnegative controllxlOOl)
                                    4010A-10                          Revision 1
                                                                    January 1995

-------
                                     TABLE 5
                    EFFECT OF WATER  CONTENT IN SOIL SAMPLES3
Soil % Water Fortified?
Mean  Std. Dev.   ± 2 SD Range
SI
SI
SI
SI
S2
S2
S2
S2
S3
S3
S3
S3
0
30
0
30
0
30
0
30
0
30
0
30
No
No
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
104.5*
101.9
38.9
49.2
97.8
105.1
40.2
48.8
98.3
111.5
43.3
46.5
106.5
106.3
47.2
51.1
105.7
109.7
47.5
47.2
107.1
103.1
47.2
49.8
99.7
95.2
40.2
48.2
96.7
93.9
42.7
44.8
99.7
95.1
43.2
48.0
103.6
101.1
42.1
49.5
100.1
102.9
43.5
46.9
101.7
103.2
44.6
48.1
3.5
5.6
4.4
1.5
4.9
8.1
3.7
2.0
4.7
8.2
2.3
1.7
96.6
89.9
33.3
46.5
90.3
86.7
36.1
42.9
92.3
86.8
40.0
44.7
- Ill
- 112
-50.9
- 52.5
- 110
- 119
- 50.9
- 50.9
- Ill
- 120
- 49.2
- 51.5
* All values  shown  are %Bo [= (ODsample/ODneaatlve Control)xl00]

a EnviroGard™ PCP in Soil  (Millipore,  Inc.)
                                    4010A-11
                     Revision  1
                  January  1995

-------
                                   TABLE 6
                        EFFECT OF pH OF SOIL  SAMPLES3
oil pH Ad.i. Fortified? Rep.
SI
SI
SI
SI
SI
SI
S2
S2
S2
S2
S2
S2
S3
S3
S3
S3
S3
S3
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
103.
88.
85.
52.
57.
44.
105.
104.
93.
47.
51.
43.
92.
96.
87.
55.
55.
44.
1* Rep. 2
1
7
2
7
1
6
6
4
4
8
4
3
3
6
7
2
3
3
98.6
96.9
90.9
44.8
44.6
41.6
93.9
91.3
87.7
45.1
44.4
40.7
101.8
91.9
99.8
49.5
48.3
39.3
Rep. 3
98.6
100.2
98.0
45.8
45.2
45.9
102.5
105.8
105.8
44.3
54.1
44.0
100.4
98.5
96.3
55.9
42.0
48.0
Mean Std. Dev. +2
100.1
95.3
91.3
47.8
48.9
44.0
100.7
100.5
95.6
45.7
50.0
42.7
98.2
95.7
94.6
53.6
48.5
43.9
2.6
5.9
6.4
4.3
7.0
2.2
6.1
8.0
9.3
1.8
5.0
1.8
5.2
3.4
6.2
3.5
6.7
4.4
94
83
78
39
34
39
88
84
77
42
40
39
87
88
82
46
35
35
SD Ranqe
.9 -
.5 -
.5 -
.2 -
.9 -
.6 -
.5 -
.5 -
.0 -
.1 -
.0 -
.1 -
.8 -
.9 -
.2 -
.6 -
.1 -
.1 -
105
107
104
56.4
62.9
48.4
113
117
114
49.3
60.0
46.3
109
103
107
60.6
61.9
52.7
All values  shown are %Bo  [=  (ODsample/ODnegatlve C0ntrol)xl00]

EnviroGard™  PCP  in  Soil (Millipore,  Inc.)
                                   4010A-12                           Revision 1
                                                                   January 1995

-------
                                    TABLE 7
                                Test Kit3 Bias
Soi1# Fortiflcation(ppm)   Extraction^ Recovered(ppm)*   % Recovery

SI            10                  19              91
SI            10                  29              86
SI            10                  39              88
SI           100                  1              84              84
SI           100                  2              78              78
SI           100                  3              76              76


S2            10                  1              10             100
S2            10                  28              76
S2            10                  38              76
S2           100                  1             101             101
S2           100                  2              98              98
S2           100                  3              88              88


S3            10                  17              72
S3            10                  28              76
S3            10                  38              81
S3           100                  1              95              95
S3           100                  2              90              90
S3           100                  3              87              87
Average »»»»»»»>»»»»»»»»»»»»»»»»»»»   84
                        Overall Average %Recovery = 86
  EnviroGard™ PCP in Soil  (Millipore,  Inc.)
                                   4010A-13                         Revision 1
                                                                  January 1995

-------
                                     TABLE 8
                  Effect  of Co-contamination with Diesel  Oila
Soi1#  Adulterated   Fortified    Rep.fll
SI
SI
SI
SI
S2
S2
S2
S2
S3
S3
S3
S3
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
YES
Rep,
  Figures are %Bo  =[(ODsample/ODnegatlve control)*100]

  EnviroGard™ PCP  in Soil  (Millipore,  Inc.)
Mean
NO
NO
YES
YES
NO
NO
YES
YES
NO
NO
YES
YES
103.2*
93.4
52.7
50.9
103.1
85.4
47.8
44.6
98.9
103.8
55.2
50.4
92.5
99.4
44.8
49.7
98.3
95.1
45.1
50.8
95.4
99.7
49.5
50.6
99.8
106.2
45.8
44.6
102.3
99.9
44.3
49.0
108.1
101.4
55.9
56.7
98.5
99.7
47.8
48.4
101.2
93.5
45.7
48.1
100.8
101.6
53.6
52.6
                                    4010A-14
                         Revision  1
                       January  1995

-------
                                    TABLE  9
                               Field Duplicates3


                                                  Sample  ID

                                   059       073       074       086       087

Method 8270

             Determination  #1     9600      74.8    836        6.59    34.0

             Determination  #2    10300      78.2   1520        6.88    51.8

                      Average     9950      76.5   1178        6.74    42.9

           Standard  Deviation      495       2.4    484        0.20    12.6

   % Coefficient  of  Variation        5.0     3.1     41        3.0     29



Immunoassay*

             Determination  #1     4480      79.5    604        2.4     36.0

             Determination  #2     3370     122      421        5.0     24.0

                      Average     3920     101      512        3.7     30.0

           Standard  Deviation      785      30.0    129        1.8      8.5

   % Coefficient  of  Variation       20      30       25       50       28

* For the purpose of this comparison, quantitative values were calculated for
the immunoassay.

a  EnviroGard™ PCP in Soil  (Millipore,  Inc.)
                                   4010A-15                         Revision 1
                                                                  January 1995

-------
                                  TABLE  10
                    Immunoassay3 Compared to Method 8270
                     Test Interpretation at 10 ppm PCP                             ^^


      Sample ID        Method 8270       Immunoassay      Concurrence?

        059               9600               >10               YES
        059D             10300               >10               YES
        060               1010               >10               YES
        061               2740               >10               YES
        063               1610               >10               YES

        064               1980               >10               YES
        065               1580               >10               YES
        066                 57.8             >10               YES
        067                110               >10               YES
        068                 47.7             >10               YES

        069                798               >10               YES
        070               2890               >10               YES
        071                289               >10               YES
        072                326               >10               YES
        073                 74.8             >10               YES

        073D                78.2             >10               YES
        074                836               >10               YES
        074D              1520               >10               YES
        075               3690               >10               YES
        076               4590               >10               YES

        077               2040               >10               YES
        078               1720               >10               YES
        079                792               >10               YES
        080               2550               >10               YES
        081                125               >10               YES

        082               2400               >10               YES
        083                270               >10               YES
        084               1140               >10               YES
        085                 57.7             >10               YES
        086                  6.59             <10               YES

        086D                 6.88             <10               YES
        087                 34.0             >10               YES
        087D                51.8             >10               YES

EnviroGard™ PCP in Soil (Millipore,  Inc.)
                                 4010A-16                         Revision 1
                                                                January  1995

-------
                      TABLE  10  (continued)
              Immunoassay3 Compared to Method 8270
              Test  Interpretation at  100 ppm PCP
Sample ID        Method 8270       Immunoassay       Concurrence?

  059               9600               >100               YES
  059D             10300               >100               YES
  060               1010               >100               YES
  061               2740               >100               YES
  063               1610               >100               YES
  064               1980               >100               YES
  065               1580               >100               YES
  066                 57.8             <100               YES
  067                110               >100               YES
  068                 47.7             <100               YES
  069                798               >100               YES
  070               2890               >100               YES
  071                289               >100               YES
  072                326               >100               YES
  073                 74.8             <100               YES
  073D                78.2             >100          False Positive
  074                836               >100               YES
  074D              1520               >100               YES
  075               3690               >100               YES
  076               4590               >100               YES
  077               2040               >100               YES
  078               1720               >100               YES
  079                792               >100               YES
  080               2550               >100               YES
  081                125               <100          False Negative
  082               2400               >100               YES
  083                270               >100               YES
  084               1140               >100               YES
  085                 57.7             <100               YES
  086                  6.59            <100               YES
  086D                 6.88            <100               YES
  087                 34.0             <100               YES
  087D                51.8             <100               YES
                           4010A-17                         Revision  1
                                                          January 1995

-------
                                  METHOD 4015

         SCREENING FOR 2,4-DICHLORORPHENOXYACETIC ACID BY IMMUNOASSAY


1.0   SCOPE AND APPLICATION

      1.1   Method 4015 is a procedure for  screening  soils and aqueous matrices
to determine whether  2,4-dichlorophenoxyacetic acid (2,4-D)  (CAS  Registry 94-75-
7) is likely to be  present  at concentrations above 0.1, 0.5,  1.0  or  5.0 mg/kg in
soil,  and   in  aqueous matrices  above  10  mg/L  (the  toxicity  characteristic
regulatory  action  level)  and  10  /jg/L  (ground water  monitoring).   Method 4015
provides  an estimate  for  the concentration  of  2,4-D by  comparison  against
standards.

      1.2   Using  the  test kit from which  this method was  developed,  >95% of
aqueous samples confirmed to have concentrations of 2,4-D below detection limits
will produce a negative result in the 10 ppm test configuration.

      1.3   In  cases  where  the  exact  concentration of  2,4-D is  required,
additional techniques  (i.e., gas chromatography Method 8151) should be used.

      1.4   This method  is restricted to  use  by  or  under  the  supervision of
trained  analysts.    Each   analyst  must  demonstrate   the  ability  to  generate
acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Test  kits  are  commercially  available   for  this  method.    The
manufacturer's directions  should be followed.

      2.2   In  general,  the method  is  performed  using an  extract of  a soil
sample, or directly on  an aqueous sample. Filtered extracts  may  be  stored cold,
in the dark.   An aliquot of the  aqueous  sample  or extract  and  an enzyme-2,4-D
conjugate reagent  are  added  to immobilized 2,4-D antibody.  The enzyme-2,4-D
conjugate  "competes"  with  2,4-D  present in  the  sample  for binding  to  2,4-D
antibody.  The  enzyme-2,4-D conjugate bound  to  the  2,4-D antibody then catalyzes
a colorless substrate to a colored product.  The test is interpreted  by comparing
the color produced  by a sample to the response  produced by a  reference reaction.


3.0   INTERFERENCES

      3.1   Compounds  that  are chemically  similar  may cause a  positive test
(false positive) for  2,4-D. The  data  for the lower limit of detection of these
compounds are provided  in Tables 1A and 1C.  Consult the information provided by
the manufacturer of  the  kit used  for additional  information  regarding  cross
reactivity with other compounds.
                                    4015-1                          Revision 0
                                                                  January 1995

-------
            3.1.1 Solutions of  Silvex  alone,  and Silvex/2,4-D  mixtures,  were
      prepared  in  TCLP  buffer  to  demonstrate  the  potential  effect  of  a
      structurally  similar, environmentally significant  cross-reactant on the
      immunoassay screening results.  At  one-half  of  the action level for 2,4-D
      (5ppm), 200 ppm of Silvex are required to be present to generate a false
      positive response.  These results are summarized in Table IB.

      3.2   Storage  and  use temperatures  may  modify the  method  performance.
Follow the manufacturer's directions for storage and use.


4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay test kit: 2,4-D RaPID,, Assay  kit  (Ohmicron),  EnviroGardIH
2,4-D in Soil  (Millipore, Inc.), or  equivalent.  Each  commercially available test
kit will supply or specify the  apparatus  and materials necessary for successful
completion of the test.


5.0   REAGENTS

      5.1   Each  commercially  available  test kit will  supply or  specify the
reagents necessary for successful  completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory material  to this  chapter,  Organic Analytes,
Section 4.1.

      6.2   Soil samples may be contaminated, and should therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions for the  test kit being used.
Those test kits  used must meet or exceed the performance specifications indicated
in Tables 2-9.
8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit being used
for quality control  procedures  specific to the test  kit  used.   Additionally,
guidance provided in Method 4000 and Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly  when  results  indicate
concentrations  near  the  action  level,   is  recommended to  refine  information
gathered with the kit.


                                    4015-2                          Revision 0
                                                                  January 1995

-------
      8.3   Do not use test kits past their expiration date.

      8.4   Do not use tubes or reagents designated for use with other test kits.

      8.5   Use  the  test kits within their  specified  storage temperature  and
operating temperature limits.

      8.6   Method 4015 is intended for field or laboratory use.  The appropriate
level of quality assurance  should  accompany the  application of this method  to
document data quality.


9.0   METHOD PERFORMANCE

      9.1   Sensitivity of the EnviroGardra 2,4-D in Soil Test Kit was determined
by establishing  the  "noise"  level  expected from  matrix effects encountered  in
negative soil samples and determining the corresponding 2,4-D concentration  by
comparison to the analyte-specific response curve.   Eight different soils  which
did  not  contain  2,4-D  were   assayed.   Each of  these  soils  was  extracted  in
triplicate and each extract was assayed  in three different assays.   The  mean  and
the standard deviation of the  resulting %Bo's (%Bo = [ (ODsample/ODnegatlve Control)xl00])
were calculated  and  the sensitivity was estimated at  two  standard deviations
below the mean.   The sensitivity  for Method 4015  was determined  to  be 80%  Bo  at
a 95% confidence  interval.  Based on the average assay response to  2,4-D, this
corresponds to 0.16 ppm  2,4-D.  These data are shown in Table 2.

      9.2   The  effect of water  content of the soil samples was determined  by
assaying three different soil  samples which had been dried and subsequently  had
water added to 30%  (w/w).  Aliquots  of  these samples were  then fortified with
2,4-D.    Each  soil  sample was  assayed three times, with  and without  added water,
and with and without 2,4-D fortification.   It was determined that water in soil
up to 30% had  no  detectable effect on the method.   These data are shown  in  Table
3.

      9.3   The effect of the  pH of the soil extract was determined  by adjusting
the soil  pH of three  soil  samples.  Soil samples were adjusted to pH 2  - 4  using
6N HC1  and pH 10  - 12 using 6N NaOH.  Aliquots of  the pH adjusted  soil samples
were fortified with 2,4-D.  Each  soil  sample was  assayed unadjusted and with  pH
adjusted to 2-4  and  10-12, both  unfortified  and  fortified.   It was determined
that soil samples with pH ranging from 3 to 11 had no detectable effect on  the
performance of the method.  These data are shown in Table 4.

      9.4   The  method  bias  was  estimated by  fortifying three  different soil
samples at two different concentrations (0.3 and 2 ppm 2,4-D).  Each fortified
sample was  extracted three  times  and  each  extract was assayed  three times.
Recovery for individual determinations ranged from 27% to 151%.  Average  recovery
for each individual extract ranged from 70% to 120%.  Overall average  recovery
for all samples was 99.7%.  These data are summarized in Table 5.

      9.5   The  probabilities of generating  false  negative  and  false  positive


                                    4015-3                          Revision 0
                                                                  January 1995

-------
results at a 10 ppm action level are shown in Table 6.

      9.6   The results obtained from spiking 2,4-D into TCLP leachates and other
aqueous samples are  reported  in  Table 7.  Each matrix  was  diluted 1:1000 and
tested by  immunoassay  5 times.  The  results are  reported  as  positive  (+)  or
negative (-).   Municipal  water results are based on a 52  ppb  cutoff to determine
positive from negative, and were diluted 1:7.

      9.7   Comparison  of  the results from  immunoassay and GC  (Method 8150)
testing of aqueous samples are presented in Table 8.

      9.8   A  field trial  was  undertaken  to  evaluate  the  ability  of  the
EnviroGardw 2,4-D  in  Soil  Test Kit to  identify 2,4-D  contaminated  soil  at a
remediation site.    A  total   of  30 soil  samples  were  evaluated by  both  the
immunoassay and  Method  8151.    Interpretation  of  the  results  at  200  /tg/kg
resulted  in 0/32   (0%)  false  negatives  and  1/32  (3%)  false positives.   This
corresponds to specificity  95% and  sensitivity of 100%.  These data  are shown in
Table 9.
10.0  REFERENCES

1.  2,4-D RaPID, Assay kit  Users  Guide,  Ohmicron.

2.  EnviroGardM 2,4-D  in  Soil  Test  Kit  Guide,  Millipore,  Inc.

3.  Lawruk,  T.S., Hottenstein, C.S., Fleeker,  J.R.,  Hall,  J.C.,  Herzog, D.P.,
    Rubio,  P.M., "Quantitation of 2,4-D  and Related Chlorophenoxy Herbicides by
    A  Magnetic   Particle-Based   ELISA"'  1993,   (manuscript   submitted  for
    publication).

4.  Hayes,  M.C., Jourdan,  S.W.,  Lawruk,  T.S.,  and  Herzog,  D.P.,  "Screening of
    TCLP Extracts of Soil  and  Wastewater for 2,4-D by  Immunoassay", USEPA Ninth
    Annual  Waste Testing and Quality Assurance Symposium, 1993.
                                    4015-4                          Revision 0
                                                                  January 1995

-------
TABLE 1A
Cross-Reactivity3 of Chi orophenoxy Compounds and
Structurally Unrelated Pesticides
Compound
2,4-D
2,4-D propylene glycol ester
2,4-D ethyl ester
2,4-D isopropyl
2,4-D methyl ester
2,4-D sec-butyl ester
2,4-D butyl ester
2,4-D butoxyethyl ester
2,4,5-T methyl ester
2,4-D iso-octyl ester
2,4-D butoxy-propylene ester
2,4-DB
MCPA
2,4,5-T
Silvex methyl ester
4-Chlorophenoxyacetic acid
MCPB
Silvex (2,4,5-TP)
Dichlorophenol
Dichloroprop
Triclopyr
MCPP
Mecoprop
Pentachlorophenol
Picloram
Concentration Giving
a Positive Result
(ppm TCLP Leachate)
10
0.52
0.54
0.96
1.09
1.40
1.60
2.00
12.0
20.0
20.6
95
110
130
665
815
980
1375
2380
5000
>10,000
>10,000
>10,000
>10,000
>10,000
Percent Cross-
Reactivity
100
1900
1850
1040
917
714
625
500
86
50
49
11
9
8
1.5
1.2
1.0
0.7
0.4
0.2
<0.1
<0.1
<0.1
<0.1
<0.1
4015-5
  Revision 0
January 1995

-------
                                TABLE 1A
            Cross-Reactivity3 of Chlorophenoxy Compounds and
                    Structurally Unrelated Pesticides
           Compound
Alachlor
Aldicarb
Aldicarb sulfate
Aldicarb sulfoxide
Atrazine
Benomyl
Butyl ate
Captan
Captofol
Carbaryl
Carbofuran
Dicamba
1,3-Dichloropropene
Dinoseb
Metolachlor
Metribuzin
Simazine
Terbufos
Thiabendazol
Concentration Giving
  a  Positive  Result
 (ppm TCLP  Leachate)
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
    >10,000
Percent Cross-
  Reactivity
 a 2,4-D RaPID^ Assay kit
TABLE IB
Cross-Reactivity3 of 2,4-D with Silvex
Si 1 vex Concentration
(ppm)
0
0.5
1.0
2.0
100
200
0
0.5
1.0
2.0
100
200
2,4-D Concentration
(ppm)
0
0
0
0
0
0
5.0
5.0
5.0
5.0
5.0
5.0
Screening Result

_
-
-
-
-
-
-
-
-
-
-
+
 a 2,4-0 RaPIDtl( Assay kit
                                 4015-6
                                 Revision 0
                               January 1995

-------
TABLE 1C
CROSS REACTIVITY'
Compound
2,4-D Acid
2,4-D butyl ester
2,4-D Dichlorophenol
2,4-D isobutyl ester
2,4-D isopropyl ester
2,4-D methyl ester
2,4-DB
2,4-DB butyl ester
Dichloroprop
Diclofop
MCPA
2,4,5-T acid
Concentration Required for
Positive Interpretation (ppm)
0.2
0.025
1.5
0.2
0.2
0.1
0.2
0.9
6.0
42.5
0.8
7.0
EnviroGardIH 2,4-D in Soil  Test Kit (Millipore Corporation)
                                  4015-7
  Revision 0
January 1995

-------
TABLE 2
Sensitivity of the EnviroGardw 2,4-D in Soil Test Kit
Part 1 - Average Response with Negative Soils
Soil#
SI
S2
S3
S4
S5
S6
S7

Average
Soil Type
LOAM
LOAM
SAND/ LOAM
CLAY
CLAY
LOAM/SAND
SAND
LOAM

Average %Bo
(n=9)
90.0
89.6
89.3
86.3
90.0
86.9
88.8
86.9
88.5
Standard
Deviation
1.7
2.3
2.1
1.9
2.3
2.6
2.8
2.9
6.5
Part 2 - Average Response with 2,4-D Calibrators
2,4-D Calibrator
Concentration (ppm)
0
0.1
0.5
1.0
5.0
Average
Absorbance
1.442
1.186
0.776
0.600
0.301
Average %Bo
N/A
82.2
53.8
41.7
20.9
Part 3 - Method Sensitivity
Based on Part 1 and Part 2 Above:
Average %Bo - 2 SD = 75.6 which is equivalent to 0.16 ppm 2,4-D
Average %Bo - 3 SD = 69.1 which is equivalent to 0.23 ppm 2,4-D
(%Bo =  [(ODsample/ODnegatpvecontrol)xlOO])
                                       4015-8
  Revision 0
January 1995

-------
                                      TABLE 3

Effect of Water Content  of Soil Sampleson the  EnviroGard™  2,4-D in Soil Test Kit


 Soil %  Water Fortified?  Rep. 1  Rep.  2  Rep.  3   Mean  Std. Dev.   ±  2 SD Range
SI
SI
SI
SI
S2
S2
S2
S2
S3
S3
S3
S3
0
30
0
30
0
30
0
30
0
30
0
30
No
No
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
98.
96.
61.
63.
98.
96.
47.
37.
98.
97.
41.
43.
7*
0
4
1
5
0
6
6
7
3
0
1
99.9
95.4
62.0
59.9
90.7
95.4
47.0
37.7
94.1
97.2
39.3
40.4
102.9
93.7
73.1
69.4
97.8
96.8
46.0
40.0
105.2
95.9
48.8
47.4
100.5
95.0
65.5
64.1
95.7
96.1
46.9
38.4
99.4
96.8
43.1
43.6
2
1
6
4
4
0
0
1
5
0
5
3
.2
.2
.6
.8
.3
.7
.8
.3
.6
.8
.1
.5
96.1
92.6
52.3
54.5
87.1
94.7
45.3
35.8
88.2
95.2
32.9
36.6
- 105
- 97.4
- 78.7
- 73.7
- 104
- 97.5
- 48.5
- 41.0
- Ill
- 98.4
- 53.3
- 50.6
   All values  shown  are %Bo [= (ODsample/ODnegatlve Control)xl00]
                                      4015-9                           Revision  0
                                                                     January  1995

-------
                                     TABLE 4

    Effect  of pH of Soil Samples  on  the  EnviroGardTk 2,4-D in Soil Test  Kit



Soil pH Adj.  Fortified?  Rep.  1*  Rep.  2  Rep. 3   Mean   Std.  Dev.  ±  2  SD Range
SI
SI
SI
SI
SI
SI
S2
S2
S2
S2
S2
S2
S3
S3
S3
S3
S3
S3
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
95
102
96
46
50
43
94
91
89
50
56
46
82
95
86
52
55
59
.5

.8
.0

.0
.3
.7
.7
.5
.3
.9
.2
.0
.1
.2
.2
.4
92.5
105
98.3
47.5
51.9
52.4
90.6
95.8
94.2
52.6
58.1
54.2
92.0
85.1
84.4
63.6
59.5
54.3
88.7
93.1
79.4
48.6
43.7
39.1
90.8
85.9
81.0
50.2
44.3
46.4
85.4
86.9
103
49.4
66.6
54.9
92.2
100
91.5
47.4
48.5
44.8
91.9
91.1
88.3
51.1
52.9
49.1
86.5
89.0
91.2
55.1
60.4
56.2
3
6
10
1
4
6
2
5
6
1
7
4
5
5
10
7
5
2
.4
.2
.5
.3
.3
.8
.1
.0
.7
.3
.5
.4
.0
.3
.4
.5
.8
.8
85.4
87.6
70.5
44.8
39.9
31.2
87.7
81.1
74.9
48.5
37.9
40.3
76.5
78.4
70.4
40.1
48.8
50.6
- 99.0
- 112
- 113
- 50.0
- 57.1
- 58.4
- 96.1
- 101
- 102
-53.7
- 67.9
- 57.9
- 96.5
- 99.6
- 112
- 70.1
- 72.0
- 61.8
* All values  shown are %Bo  [=  (ODsample/ODnegatlve control)xlOO]
                                     4015-10
  Revision 0
January  1995

-------
                                    TABLE 5

                Bias of the EnviroGard,, 2,4-D in  Soil  Test Kit
Soi1#  Fortification (ppm)   Extraction^   Recovered (ppm)* %  Recovery
SI
SI
SI
SI
SI
SI
Average >»
S2
S2
S2
S2
S2
S2
Average >»
S3
S3
S3
S3
S3
S3
Average >»
0.3
0.3
0.3
2
2
2
•»»»»»»»:
0.3
0.3
0.3
2
2
2
»»»»»»»:
0.3
0.3
0.3
2
2
2
»»»»»»»;
1
2
3
1
2
3
>»>»»»»»»:
1
2
3
1
2
3
>>»»»»»»»:
1
2
3
1
2
3
>»»»>»»»»x
0.21
0.24
0.23
1.87
2.12
2.40
»»»»»»»»:
0.29
0.29
0.30
2.05
1.89
2.22
>»»»»»>»»:
0.31
0.31
0.31
2.28
2.30
2.24
>»»>»»»»»:
70.0
80.0
76.6
93.5
106
120
»» 91.0
96.7
96.7
100
102
94.5
111
»» 100
103
103
103
114
115
112
»» 108
                       Overall Average %Recovery =99.7
                                    4015-11
  Revision 0
January 1995

-------
                                 Table 6

Probability of False Negative and False Positive Results for 2,4-D RaPIDw
  Assay kit at a  10 ppm Action Level  in TCLP Extract from Organic Soil
Spike Concentration
2,4-D (PPM)
5
7.5
10
15
Probability of False
Positive (%)
0
70
N/A
N/A
Probability of False
Negative (%)
N/A
N/A
0
0
             Results were  based on  ten  replicate spiked  samples.    Cutoff
             levels were  established using 30  replicates  of each  solution
             tested in 3 immunoassay batch runs.
             N/A  =    No  false positives  possible  above/below  the
             1imit.
      action
                                 4015-12
  Revision 0
January 1995

-------
T
2,4-D Spiking Results on
ID #
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
Matrix/Spike
TCLP Buffer
TCLP Buffer + 15 ppm
TCLP Buffer + 10 ppm
TCLP Buffer + 5 ppm
Sandy Extract13
Sandy Extract + 15 ppm
Sandy Extract + 10 ppm
Sandy Extract + 5 ppm
Organic Extract0
Organic Extract + 15
Organic Extract + 10
Organic Extract + 5
Effluent #1
Effluent #1 + 15 ppm
Effluent #1 + 10 ppm
Effluent #1+5 ppm
Effluent n
Effluent #2 + 15 ppm
Effluent #2 + 10 ppm
Effluent #2+5 ppm
Runoff
Runoff + 15 ppm
Runoff + 10 ppm
Runoff + 5 ppm
able 7
\queous Environmental Matrices3
1 	
2,4-D Test Results
Rl
-
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
_
+
+
-
-
+
+
-
R2
_
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
-
+
+
-
-
+
+
-
R3
_
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
-
+
+
-
_
+
+
-
R4
-
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
-
+
+
-
-
+
+
-
R5
_
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
_
+
+
-
_
+
+
-
%POS
-
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
_
+
+
-
-
+
+
-
%NEG
-
+
+
-
-
+
+
-
-
+
+
-
_
+
+
-
_
+
+
-
_
+
+
-
4015-13
  Revision 0
January 1995

-------
Table 7
2,4-D Spiking Results on Aqueous Environmental Matrices3

25
26
27
28
Municipal Water
Municipal Water + 140 ppb
Municipal Water + 70 ppb
Municipal Water + 35 ppb
2,4-D Test Results
_
+
+
-
_
+
+
-
_
+
+
-
_
_
_
-
_
-
_
-

_
_
_
-
_
-
_
-
                                                                    i
8 2,4-D RaPIDw Assay  kit
b Sandy Soil  TCLP Extract
c Organic  Soil  TCLP Extract
                    4015-14
  Revision 0
January 1995

-------
Table 8
2,4-D Spiking Results
2,4-D RaPIDM Assay kit vs. Method 8151
ID*
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
Matrix/Spike
TCLP Buffer
TCLP Buffer + 15 ppm
TCLP Buffer + 10 ppm
TCLP Buffer + 5 ppm
Sandy Extract3
Sandy Extract + 15 ppm
Sandy Extract + 10 ppm
Sandy Extract + 5 ppm
Organic Extract13
Organic Extract + 15 ppm
Organic Extract + 10 ppm
Organic Extract + 5 ppm
Effluent #1
Effluent #1 + 15 ppm
Effluent #1 + 10 ppm
Effluent #1+5 ppm
Effluent #2
Effluent #2 + 15 ppm
Effluent #2 + 10 ppm
LEf fluent #2+5 ppm
Runoff
Runoff + 15 ppm
Runoff + 10 ppm
Immunoassay
Results
5/5 Negative
5/5 Positive
5/5 Positive
5/5 Negative
5/5 Negative
5/5 Positive
5/5 Positive
5/5 Negative
5/5 Negative
5/5 Positive
5/5 Positive
5/5 Negative
5/5 Negative
5/5 Positive
5/5 Positive
5/5 Negative
5/5 Negative
5/5 Positive
5/5 Positive
5/5 Negative
5/5 Negative
5/5 Positive
5/5 Positive
Method 8151
2,4-D (ppm)
nd
13.0
11.0
5.6
nd
*
5.9, 5.2
*
nd
*
10.0, 9.5
*
*
*
11.0, 7.8
3.6
*
11.0
8.8, 9.5
*
nd
*
9.7, 8.6
Correlation
IA vs. GC
Yes
Yes
Yes
Yes
Yes
*
No
*
Yes
*
Yes
*
*
*
Yes
Yes
*
Yes
Yes
*
Yes
*
Yes
4015-15
  Revision 0
January 1995

-------
Table 8
2,4-0 Spiking Results
2,4-D RaPID,, Assay kit vs. Method 8151
ID#
24
25
26
27
28
Matrix/Spike
Runoff + 5 ppm
Municipal Water
Municipal Water + 140 ppb
Municipal Water + 70 ppb
Municipal Water + 35
-
Immunoassay
Results
5/5 Negative
5/5 Negative
5/5 Positive
5/5 Positive
5/5 Negative
Method 8151
2,4-D (ppm)
5.5
nd
*
58.59 (ppb)
*
Correlation
IA vs. GC
Yes
N/A
N/A
N/A
N/A
     Sandy Soil  TCLP Extract
b    Organic Soil  TCLP Extract

nd   non-detectable
N/A  Not applicable to wastewater regulatory limit
*    No analysis with Method 8150
                    4015-16
  Revision 0
January 1995

-------
                                TABLE 9


  Comparison of the EnviroGardm 2,4-D in Soil  Test Kit to Method 8151

                Interpretation of Results  at  200  jug/kg
Sample #
Method 8150 ug/kg
Immunoassay Result
Agrees?
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
<200
<200
220
<200
<200
330
<200
<200
830
<200
310
350
<200
<200
200
<200
<200
440
560
380
NEGATIVE
NEGATIVE
POSITIVE
NEGATIVE
NEGATIVE
POSITIVE
POSITIVE
NEGATIVE
POSITIVE
NEGATIVE
POSITIVE
POSITIVE
NEGATIVE
NEGATIVE
POSITIVE
NEGATIVE
NEGATIVE
POSITIVE
POSITIVE
POSITIVE
YES
YES
YES
YES
YES
YES
FALSE POSITIVE
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
                                4015-17
                                                  Revision 0
                                               January  1995

-------
21                 <200                  NEGATIVE             YES
22                 360                   POSITIVE             YES
23                 <200                  NEGATIVE             YES
24                 <200                  NEGATIVE             YES
25                 <200                  NEGATIVE             YES

26                 <200                  NEGATIVE             YES
27                 <200                  NEGATIVE             YES
28                 <200                  NEGATIVE             YES
29                 <200                  NEGATIVE             YES
30                 <200                  NEGATIVE             YES
                            4015-18                          Revision  0
                                                           January  1995

-------
                                  METHOD 4020

            SCREENING FOR POLYCHLORINATED BIPHENYLS BY IMMUNOASSAY
1.0  SCOPE AND APPLICATION

     1.1   Method 4020 is a procedure for screening  soils  and non-aqueous waste
liquids to determine when total  polychlorinated biphenyls  (PCBs)  are present at
concentrations above 5,  10 or 50 mg/kg.  Method  4020 provides an  estimate  for the
concentration of PCBs by  comparison with a standard.

     1.2   Using the test kit from which this method was developed, 95%  of soil
samples containing 0.625  ppm or less  of  PCBs will produce a negative result in
the 5 ppm test configuration.  Using another commercially available test kit, 97%
of soil  samples  containing 0.25  ppm  or less  of PCBs will  produce a negative
result in the assay  and  greater  than  99% of the samples containing 1.0 ppm or
more will produce a  positive  result.   Tables  2-5,  7, 10,  and 11 present false
positive and false negative data generated from commercially available test kits.
Using a test kit commercially available for screening non-aqueous  waste liquids,
>95% of samples containing  0.2-0.5 ppm or  less of PCB will  produce a negative
result.

     1.3   In  cases where  the  exact  concentrations  of  PCBs  are  required,
quantitative techniques  (i.e., Method 8082) should be used.

     1.4   This method  is  restricted  to use   by  or under  the  supervision  of
trained  analysts.    Each  analyst must  demonstrate the  ability  to  generate
acceptable results with this method.

2.0  SUMMARY OF METHOD

     2.1   Test  kits  are  commercially  available  for  this  method.     The
manufacturer's  directions  should  be  followed.    In general,  the method  is
performed using a sample  extract.   Sample  and  an enzyme conjugate reagent are
added to immobilized antibody.  The enzyme conjugate  "competes" with PCB present
in the sample  for  binding  to  immobilized anti-PCB antibody.   The test  is
interpreted  by  comparing the  response  produced  by testing  a  sample  to  the
response produced by testing standard(s) simultaneously.


3.0  INTERFERENCES

      3.1  Chemically similar compounds and compounds which might  be expected to
be found  in  conjunction  with PCB contamination were tested to  determine  the
concentration required to produce a positive test  result.  These  data are shown
in Tables 1A, IB,  1C, and ID.


4.0  APPARATUS AND MATERIALS

      4.1  Immunoassay test kit:  PCB  RISc™ (EnSys,  Inc.),  EnviroGard™  PCB  in
Soil  (Millipore,  Inc.),  D  TECH™  PCB test  (Strategic Diagnostics  Inc.),  PCB


                                    4020-1                           Revision  0
                                                                  January 1995

-------
RISc™ Liquid Waste  Test System (EnSys, Inc.),  or equivalent.  Each commercially
available test kit will  supply or  specify the apparatus and materials necessary
for successful completion of the test.


5.0  REAGENTS

     5.1   Each  commercially  available test  kit will  supply  or  specify  the
reagents necessary for successful  completion of the test.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1  See the  introductory material  to  this chapter,  Organic Analytes,
Section 4.1.  Also refer to  Reference 9 for the  collection and handling of non-
aqueous waste liquids.

      6.2  Samples  may  be  contaminated,  and  should  therefore  be  considered
hazardous and handled accordingly.


7.0  PROCEDURE

     7.1   Follow the manufacturer's instructions for the test kit being used.
Those test  kits used must meet  or exceed the  performance specifications indicated
in Tables 2-11.
8.0  QUALITY CONTROL

     8.1   Follow the manufacturer's instructions for the test kit being used for
quality control  procedures specific to the test kit used.  Additionally, guidance
provided in Method 4000 and Chapter One should be followed.

     8.2   Use  of  replicate  analyses,  particularly  when  results  indicate
concentrations  near  the action  level,  is  recommended  to  refine  information
gathered with the kit.

     8.3   Do not use test kits past their expiration  date.

     8.4   Do not use tubes or reagents designated for  use with other test kits.

     8.5   Use  the  test kits within  their specified  storage  temperature  and
operating temperature limits.

     8.6   Method 4020 is  intended for field or laboratory use.  The appropriate
level of quality  assurance should accompany the  application  of this  method to
document data quality.


9.0  METHOD PERFORMANCE

      9.1  A study was  conducted  with the PCB RISc™ test  kit  using fourteen


                                    4020-2                          Revision 0
                                                                  January 1995

-------
standard  soils  and  three   soil  samples  whose  PCB  concentration   had  been
established by Method 8082.   Replicates were  performed  on  seven of the standard
soils and on one of the soil  samples  for a  total of 25  separate analyses.  Each
of two  different analysts ran  the  25  analyses.   Results  indicated  that "<"
assignments are accurate with almost 99% certainty at the 50 ppm level while ">"
assignments  can  be  up to about  96%  inaccurate  as  the  sample  concentration
approaches that  of  the testing  level.   Corresponding certainties at the 5 ppm
level are 92% and 82%  respectively.   Tables 2 and 3 summarize these results.

     9.2   Table 4 presents method precision data generated using  the PCB RISc™
test kit, comparing immunoassay test  results  with  results  obtained using Method
8082.

     9.3   Method precision was determined  with  the EnviroGard PCB in Soil test
kit by assaying 4 different  soils  (previously determined to  contain 5.04, 9.78,
11.8, and  25.1  mg/kg  by  Method 8082),  at  three  different  sites,  using three
different lots  of assay  kits,  three  times a day for  9  days.    A total  of 81
analyses were performed for each soil.  Error attributable to site,  lot, date,
and operator were determined.  Separately,  the relative reactivity of Aroclors
1242, 1248,  1254, and 1260 were determined.   Based  on  Aroclor heterogeneity, and
method  imprecision,  concentrations  of  Aroclor  1248  were  selected  that would
result in greater than 99% confidence  for negative interpretation.  A study was
conducted  (Superfund  SITE   demonstration)  on  114  field  samples  whose  PCB
concentration were also determined by Method 8082.   32 of the field samples were
collected in duplicate  (as coded field duplicates) and assayed by standard and
immunoassay methods.  The results  for all 146 samples are summarized in Tables
5 and 6.

     9.4   Grab  samples were obtained  from  sites  in  Pennsylvania,  Iowa and
Illinois using a stainless steel trowel.  Each sample  was homogenized by placing
approximately six cubic inches in  a  stainless steel  bucket and mixing with the
trowel for approximately two  minutes.  The soils  was aliquotted into 2 six ounce
glass bottles.  The samples were tested on  site  using the D TECH PCB test kit,
and sent to an analytical laboratory  for analysis by Method 8082.   These data are
compared in Table 7.

     9.5   Tables 8 and 9 present  data on the inter-  and intra-assay precision
of the PCB RISc™ Liquid  Waste Test System.   The  data  were generated  using 11
samples, each spiked at 0, 0.2 and 5 ppm, and assayed 4 times.

     9.6   Tables 10  and  11  provide data  from  application of the  PCB RISc™
Liquid  Waste Test  System  to  a  series of   liquid  waste  samples  whose  PCB
concentration had been established by Method 8082.


10.0 REFERENCES

1.   J.P. Mapes, T.N. Stewart, K.D.  McKenzie, L.R. McClelland,  R.L.  Mudd,  W.B.
     Manning, W.B.  Studabaker,  and  S.B.  Friedman,  "PCB-RISc™  - An  On-Site
     Immunoassay for  Detecting  PCB  in Soil", Bull.  Environ.  Contain.  Toxicol.
     (1993)  50:219-225.

2.   PCB RISc™ Users Guide,  Ensys  Inc.


                                    4020-3                           Revision 0
                                                                  January 1995

-------
3.   R.W. Counts, R.R.  Smith, J.H. Stewart, and R.A.  Jenkins, "Evaluation of PCB
     Rapid Immunoassay Screen Test System", Oak Ridge National  Laboratory, Oak
     Ridge, TN 37831, April 1992, unpublished

4.   EnviroGard PCB in Soil Package Insert, Millipore Corp.  2/93.

5.   Technical Evaluation  Report  on  the Demonstration of  PCB  Field Screening
     Technologies, SITE Program.  EPA Contract Number 68-CO-0047.   2/93.

6.   D TECH™ PCB Users Guide ,  SDI/Em Sciences

7.   Melby, J.M., B.S. Finlin,  A.B. McQuillin, H.G.  Rovira, J.W.  Stave, "PCB
     Analysis by Enzyme Immunoassay",  Strategic Diagnostics Incorporated,
     Newark, Delaware, 1993

8.   Melby, J.M., B.S. Finlin,  A.B. McQuillin, H.G.  Rovira, "Competitive
     Enzyme Immunoassay (EIA) Field Screening System for the Detection of
     PCB", 1993 PCB Seminar, EPRI, September 1993

9.   T.A. Bellar and J.J Lichtenberg.   The Analysis  of Polychloringated
     Biphenyls in Transformer Fluid and Waste Oils.   U.S. EPA Research and
     Development, EPA/EMSL-ORD,  Cincinnati, Ohio (June 24,  1980).   Revised
     June 1981, EPA 600/4-81-045.

10.  PCB RISc™ Liquid Waste Test System,  User's Guide,  EnSys  Environmental
     Products, Inc.
                                    4020-4                          Revision 0
                                                                  January 1995

-------
TABLE 1A
CROSS REACTIVITY OF DIFFERENT COMPOUNDS"
Compound
1-Chloronaphthalene
1,2,4-Trichlorobenzene
2,4-Dichlorophenyl-benzenesulfonate
2,4-Dichloro-l-naphthol
Bifenox
Diesel fuel
Pentachlorobenzene
2,5-Dichloroanil ine
Hexachlorobenzene
Gasol ine
Dichlorofenthion
Tetradifon
Soil Equivalent Concentration (ppm)
Required to Yield a Positive Result
10,000
10,000
1,000
>10,000
500
>10,000
>10,000
>10,000
>10,000
>10,000
10,000
125
(a)   PCB RISc'M test kit, Ensys, Inc. publication
                                    4020-5
  Revision 0
January 1995

-------
TABLE IB
CROSS REACTIVITY OF DIFFERENT COMPOUNDS'
Compound
Aroclor 1248
Aroclor 1242
Aroclor 1254
Aroclor 1260
1,2-, 1,3-, & 1,4-Dichlorobenzene
1,2,4-Trichlorobenzene
biphenyl
2,4-dichlorophenol
2,5-dichlorophenol
2,4,5-trichlorophenol
2,4,6-trichlorophenol
Pentachlorophenol
% Cross Reactivity
100
50
90
50
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
" EnviroGard PCB Test Kits (Millipore Corporation)
                                    4020-6
  Revision 0
January 1995

-------
TABLE 1C
CROSS REACTIVITY OF DIFFERENT COMPOUNDS"
Compound
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1260
Aroclor 1262
Aroclor 1268
MDLb
(ppm)
5.7
25.5
9.0
1.5
0.8
0.5
0.75
0.5
3.8
1C 50C
(ppm)
83
300
105
31
24
10
10
10
40
% CROSS REACTIVITYd
12
3
10
32
42
100
100
100
25
METHOD:       The compounds listed were assayed at various concentrations and
              compared against an inhibition curve generated using Aroclor
              1254.  The concentration of the compound required to elicit a
              positive response at the MDL as well as the concentration
              required to yield 50% inhibition compared to the standard curve
              were determined.
   D TECH'M PCB test kit
   The IC50  is defined as  the  concentration of compound  required  to  produce  a
test response equivalent  to 50% of the maximum response.


c  The Minimum Detection Limit (MDL)  is defined  as  the lowest concentration  of
compound that yields a positive test result.
d  % Crossreactivity is determined by dividing the equivalent Aroclor 1254
concentration by the actual compound concentration at  IC50
                                    4020-7
  Revision 0
January 1995

-------
TABLE ID
CROSS REACTIVITY OF DIFFERENT COMPOUNDS'
Compound
1- Chi oron aphtha! ene
1,2,4-Trichlorobenzene
2,4-Dichloro-l-naphthol
Bifenox
Pentachl orobenzene
2,5-Dichloroanil ine
Hexachlorobenzene
Dichlorofenthion
Tetradifon
% Cross-Reactivity
0.05%
0 . 05%
<0.20%
<0.10%
<0.05%
<0.05%
<0.05%
0.05%
<0.10%
Soil Equivalent
Concentration (ppm) Required
to Yield a Positive Result
10,000
10,000
>10,000
500
>10,000
>10,000
>10,000
10,000
125
(a)     PCB  RISc'M  Liquid  Waste  Test  System,  Ensys,  Inc.
                                    4020-8
  Revision 0
January 1995

-------








ro
O
»-H
H-
_l
i— i
O
Q.
in
o
u.

CO
UJ
i—
<
o£

o
DC:
o£

o
UJ
fe

CO
UJ









o
CVJ

o

cn

CO
i^
10
in

•**•


CO




cvj

i — i


o
'E"
Q.
Q.
QJ
13
"~fa
>
OJ
3
S-
1—







•
•
•
•
CO
CVJ
CO
CVJ
in
UO


CVJ
cn
CO
CVJ
CO

CO
i— (
a>
+j - — -
n3 55
^ a) — '
-0 ^ oo
OJ "~Z 01
s-z>
E o_ -i-1
"^ o'S
00 O
UJ Q.
1 — 1
o
•v
CO
O
LO
o
o
o
CVJ
*d-
LO
co

•


•






•



a>
+-> - — -
fO 5?
^ QJ —^
-0 ^ oo
0)^(1)
*2*
E M +->
"-^ «
+-> ° cn
00 QJ
UJ Z.


(D
O
1— 1
h-
_l
o
o.
o.
0
in
C£.
o
u.
CO
UJ
i—
<
o;
C£
0
ct:
ac
UJ
o
UJ
<
i-^
i—
CO
UJ




o

o
co
o
1 —
o


o
m

o

CD
CO

CVJ
IT)
1 — 1
o

in

CD

a>
13
fC """"^
> E
CL
OJ Q.
S_
1—


•
•





UD
m
cn
CO
r^
co
O
LO
l£)
o
l£>
^1-
LD
CVI
cn
r--.
o
cu
TJ 55
^ n> ^^
-a °° oo
|S|
E, -,
M- ^
'-^ ° 'o^
00 O
UJ Q_
—
O
\/
CVJ
0
CO
CD
r-~
o

r^
i— i






•





a>
03 55
^ QJ ~—
-0 ^ c^
5^>
== M- -tl
•^ ° s.
oo a>
LU 2:
        O LD
            CT)
        -r-  S_
         >  ft)
         a>  3
        o:  c
             rd
         CD
         CNJ
         O
 a>
 o
oo
CO
(_J
Q-

-------
Table 4
Comparison of PCB RISc Test Kit with GC
Sample 10
101
284
292
199
264
257
259
265
200
170
198
172
169
171
202
163
165
168
166
164
204
253
203
258
106
161
167
Screening Test
Results
<5 ppm
<5 ppm
<5 ppm
<5 ppm
<5 ppm
<5 ppm
<5 ppm
<5 ppm
<5 ppm
5-50
<5 ppm
5-50
5-50
5-50
<5 ppm, 5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
GC Results
<0.5 ppm
<0.5 ppm
<0.5 ppm
0.5 ppm
1 ppm
1.8 ppm
4 ppm
4.5 ppm
5 ppm
5.8 ppm
2.2-5.8 ppm
6.2 ppm
7.2 ppm
7.2 ppm
1.3-7.2 ppm
8.7 ppm
9 ppm
9 ppm
9.3 ppm
11.9 ppm
12.8 ppm
13 ppm
13.5 ppm
15 ppm
15-19 ppm
15.3 ppm
16.2 ppm
AGREEMENT8
Y, FP, FN
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
4020-10
  Revision 0
January 1995

-------
Table 4
Comparison of PCB RISc Test Kit with GC
Sample 10
247
148
205
162
175
176
197
243
252
178
201
254
238
248
250
242
256
249
245
241
246
261
240
267
239
104
108
Screening Test
Results
5-50
>50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50
5-50, >50
>50
5-50
>50
5-50
>50
>50
>50
5-50
>50
>50
>50
>50
>50
>50
>50
GC Results
18 ppm
18-34 ppm
20 ppm
20.4 ppm
21.2 ppm
21.6 ppm
32 ppm
32 ppm
32 ppm
43.7 ppm
43 ppm
56 ppm
46-60 ppm
44-60 ppm
68 ppm
30-69 ppm
73 ppm
96 ppm
102 ppm
154 ppm
154 ppm
204 ppm
251 ppm
339 ppm
460 ppm
200-3772 ppm
531-1450 ppm
AGREEMENT"
Y, FP, FN
Y
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FN
Y
Y
Y
Y
Y
Y
Y
Y=Yes, FN=False Negative, FP=False Positive

                               4020-11
  Revision 0
January 1995

-------
                Table 5

Comparison of EnviroGard PCB Kit with GC
SAMPLE
NUMBER
001
002
003
004
005
006
007
008
009
010
Oil
012
013
014
015
015D
016
017
018
019
020
021
022
022D
023
024
SCREENING GC RESULT0
RESULTc'd [8082]
>10 5.98
>10 1.27
<10 0.11
>10 6.71
>10 1.37
>10 0.68
>10 0.55
>10 2.00
>10 1.30
>10 0.17
>10 1.15
<10 NDf
<10 1.13
<10 0.18
>10 9.13
>10 9.84
>10 2110
>10 2.55
>10 45.4
>10 6.70
<10 0.07
<10 0.06
<10 0.54
<10 0.72
>10 20.8
<10 0.06
AGREEMENT
Y, FN, FP
ppg
FP
Y
ppg
FP
FP
FP
FP
FP
FP
FP
Y
Y
Y
FPa
ppg
Y
FP
Y
ppg
Y
Y
Y
Y
Y
Y
                 4020-12
  Revision 0
January 1995

-------
Table 5 (continued)
SAMPLE
NUMBER
024D
025
026
027
028
028D
029
030
031
032
033
034
035
035D
036
037
037D
038
039
040
041
042
042D
043
043D
044
SCREENING GC RESULT0
RESULT" [8082]
<10 0.05
>10 11.7
<10 1.96
<10 0.06
<10 0.22
<10 0.22
<10 0.23
<10 1.15
<10 0.26
>10 47.6
>10 6.00
>10 34.0
<10 NDf
<10 NDf
>10 816
<10 0.06
<10 0.04
>10 1030
<10 0.68
>10 4.25
<10 ND'
>10 0.52
>10 0.47
>10 1.69
>10 1.74
<10 0.59
AGREEMENT
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP9
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
FP
FP
FP
FP
Y
      4020-13
  Revision 0
January 1995

-------
Table 5 (continued)
SAMPLE
NUMBER
045
046
046D
047
047D
048
049
050
050D
051
052
053
054
055
056
057
058
059
060
060D
061
062
063
063D
064
065
SCREENING GC RESULT0
RESULTc'd [8082]
<10 NDf
<10 NDf
<10 NDf
<10 0.09
<10 0.10
<10 NDd
<10 NDd
>10 3.60
>10 4.41
<10 NDf
>10 4.21
<10 0.96
<10 0.52
<10 2.40
<10 0.51
<10 NDf
<10 0.69
>10 7.86
>10 0.62
<10 0.58
>10 580
>10 2.35
<10 0.09
<10 0.15
>10 19.0
>10 3.08
AGREEMENT
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
FP
FP
Y
FP
Y
Y
Y
Y
Y
Y
FP°
FP
Y
Y
FP
Y
Y
Y
FP
     4020-14
  Revision 0
January 1995

-------
Table 5 (continued)
SAMPLE
NUMBER
066
067
068
069
069D
070
071
071D
072
073
074
075
076
077
078
079
080
081
081D
082
082D
083
083D
084
084D
085
SCREENING GC RESULT0
RESULT°'d [8082]
<10 1.98
<10 0.08
<10 0.50
<10 NDf
<10 NDf
<10 NDf
<10 0.05
<10 NDf
<10 0.04
>10 15.8
>10 13.3
>10 23.0
>10 46.7
<10 NDf
>10 2.27
>10 42.8
<10 3.77
<10 0.69
<10 0.45
<10 NDf
<10 0.24
<10 0.48
<10 0.41
>10 1.16
>10 1.08
>10 428
AGREEMENT
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
Y
Y
Y
Y
Y
Y
Y
FP
FP
Y
      4020-15
  Revision 0
January 1995

-------
Table 5 (continued;
SAMPLE
NUMBER
085D
086
086D
087
087D
088
088D
089
090
090D
091
091D
092
092D
093
094
095
095D
096
097
097D
098
098D
099
100
100D
SCREENING GC RESULT0
RESULTc'd [8082]
>10 465
<10 1.42
<10 1.25
<10 0.08
<10 NDf
>10 2.70
>10 1.77
>10 45.0
<10 1.01
<10 1.40
>10 1630
>10 1704
<10 1.21
<10 NDf
<10 0.30
<10 0.36
>10 17.5
>10 31.2
<10 0.06
<10 1.23
<10 0.29
>10 1.17
>10 0.83
<10 NDf
>10 177
>10 167
AGREEMENT
Y, FN, FP
Y
Y
Y
Y
Y
FP
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
FP
Y
Y
Y
     4020-16
  Revision 0
January 1995

-------
                              Table 5 (continued)
SAMPLE
NUMBER
101
102
102D
103
104
105
106
107
108
109
109D
110
111
112
113
114
SCREENING GC RESULT0
RESULT0'" [8082]
>10 1.21
>10 293
>10 177
>10 40.3
>10 7.66
<10 0.21
<10 2.50
>10 14.1
>10 3.84
<10 NDf
<10 NDf
<10 NDf
<10 NDf
>10 315
>10 14.9
>10 66.3
AGREEMENT6
Y, FN, FP
FP
Y
Y
Y
ppa
Y
Y
Y
FP
Y
Y
Y
Y
Y
Y
Y
c  mg/kg  (ppm)
d  Screening Calibrator is 5 mg/kg Aroclor 1248
e  Y=Yes,  FN=False Negative, FP=False Positive
f   ND  = Not  Detectable
8  Expected  Result Based on Calibrator Concentration
                                    4020-17
  Revision 0
January 1995

-------
                                    Table  6

                 EnviroGard PCB Kit Field Performance Summary
Specificity:    [l-(Reported Positives/True Negatives)]  =  [l-(37/109)]  = 66%
Note 1:         8 of the  37 reported  positive  samples had  PCB contamination
                levels  between  5  and  10 mg/kg.  Soils in this range should test
                "positive"  because the assay calibrator is  5 mg/kg Aroclor 1248.
                A positive assay bias  is necessary to  prevent  false negative
                results.
                Eliminating  these  samples  from  the  calculations  produces  a
                Specificity  of:
                [l-(Reported Positives/True  Negatives)]  =  [l-(29/101)]  = 71%
Note 2:         The distribution of false positives  is  not  random (p < 0.05),
                with a clustering  at  the beginning  of  the  sample  set.   This
                observation was  included in Developers Comments which  were added
                to the final  draft  of the Technical  Evaluation  Report2.   One
                explanation for  the higher frequency of false positive results
                at  the  beginning  is  inexperience  of  the  operator  with  the
                method.    If  the first  20  samples  are  eliminated  from  the
                Specificity analysis,  the following  result is  obtained:
                [l-(Reported Positives/True  Negatives)]  =  [l-(20/86)]  = 77%
                In  the  SITE  demonstration,   the  PCB  Immunoassay  had   a  77%
                positive predictive value.
Sensitivity:    [l-(Reported Negatives/True  Positives)]  =   [1-(0/31)]  =  100%
                In the  SITE  demonstration,  the  PCB  Immunoassay  had  a  100%
                negative predictive value.
                                    4020-18                         Revision 0
                                                                  January 1995

-------
                               TABLE 7

         COMPARISON  OF  D  TECH  PCB  test  kit  WITH  GC  -  TRIAL  #1
SAMPLE
01
02
J3
J5
J6
07
J8
J9
J10
Jll
J12
J13
J14
J15
016
J17
J18
J19
J20
J21
J22
J23
J24
D TECH
(ppm)
4.0-15
>50
15-50
15-50
>50
4.0-15
4.0-15
>50
>50
>50
15-50
>50
>50
15-50
15-50
>50
>50
>50
>50
>50
1.0 .
1.0
<0.5
GC
(ppm)
5.0
147
54
160
1200
12
28
463
1760
28
17
1300
186
31
36
31
130
1310
2620
11100
0.01
0.60
0.10
AGREEMENT
Y, FN, FP
Y
Y
Y
FN
Y
Y
FN
Y
Y
FP
Y
Y
Y
Y
Y
FP
Y
Y
Y
Y
FP
Y
Y
SAMPLE
J25
J26
J28
J28
J29
J30
J31
J32
J33
J34
035
036
037
038
039
040
041
042
043
044
045
046
047
D TECH
(ppm)
0.5
<0.5
1.0
<0.5
0.5
>50
4.0-15
0.5
0.5
1.0
1.0
>50
<0.5
0.5
0.5
<0.5
<0.5
1.0
1.0
15-50
15-50
<0.5
<0.5
GC
(ppm)
0.12
0.01
1.8
0.18
0.54
21
13
0.72
0.32
0.36
0.26
70
0.12
0.81
0.33
0.19
0.01
0.43
0.31
503.4
5.6
0.02
0.22
AGREEMENT
Y, FN, FP
FP
Y
Y
Y
Y
FP
Y
Y
Y
FP
FP
Y
Y
Y
Y
Y
Y
FP
FP
FN
FP
Y
Y
Y=Yes, FN=False Negative, FP=False Positive
                               4020-19
  Revision 0
Oanuary 1995

-------
                    TABLE  7(cont)

COMPARISON OF D TECH PCB test kit WITH GC - Trial #2
SAMPLE
Gl
G2
G3
G4
G5
G6
G7
G8
G9
G10
Gil
G12
G13
G14
G15
G16
G17
G18
G19
G20
D TECH
(ppm)
15-50
4.0-15
1.0-4.0
15-50
<0.5
1.0-4.0
1.0-4.0
15-50
4.0-15
15-50
4.0-15
4.0-15
4.0-15
0.5-1.0
<0.5
1.0-4.0
4.0-15
4.0-15
1.0-4.0
>50
GC
(ppm)
18
11
3.4
6.5
0.01
1.4
0.30
7.5
33
8
11
24
4.3
1.3
0.01
3.2
18
4.6
2.3
37
AGREEMENT
Y, FN, FP
Y
Y
Y
FP
Y
Y
FP
FP
FN
FP
Y
FN
Y
Y
Y
Y
Y
Y
Y
FP
                       4020-20
  Revision 0
January 1995

-------
                    TABLE  7(cont)

COMPARISON OF D TECH PCB test kit WITH GC - Trial #3
SAMPLE
W1A
W2A
W3A
W4A
W5A
W6A
W7A
W8A
W9A
W10A
W11A
W12A
W13A
W14A
W15A
W16A
W17A
W18A
W19A
W20A
W21A
W22A
W23A
D TECH
(ppm)
4.0-15
4.0-15
1.0-4.0
4.0-15
>50
>50
>50
4.0-15
1.0-4.0
0.5-1.0
15-50
15-50
15-50
4.0-15
1.0-4.0
1.0-4.0
4.0-15
1.0-4.0
4.0-15
>50
>50
1.0-4.0
>50
GC
(ppm)
9.1
11
2.8
13
29
1200
57
18
1.3
0.44
120
48
19
2.7
1.3
0.3
1.4
2.2
8.2
9.3
110
0.6
46
AGREEMENT
Y, FN, FP
Y
Y
Y
Y
FP
Y
Y
Y
Y
Y
FN
Y
Y
Y
Y
FP
FP
Y
Y
FP
Y
Y
Y
                      4020-21
  Revision 0
January 1995

-------
Table 8
Intraassay Precision of the PCB RISc™ Liquid Waste Test System
PCB 1248 Spike
Concentration
(ppm)
0
0.2
5
Signal %RSD
(OD450nm) N=44
(11 data sets)
6.4%
5.9%
7.9%
Statistical Percentage of
False Results Compared to
Standards
<0.02%
4.1%
1.4%
Table 9
Interassay Precision of the PCB RISc™ Liquid
Waste Test System
PCB 1248 Spike
Concentration (ppm)
0
0.2
5
Signal %RSD
(OD450nm) N=44
(11 data sets)
6.4%
8.3%
8.5%
4020-22
  Revision 0
January 1995

-------
                         Table 10
Comparison of PCB RISc™ Liquid Waste Test with Method 8082
Sample
ID
302
303
304
306
307
308
310
311
331
380
381
382
383
384
385
387
388
389
390
391
394
395
396
398
399
400
401
402
403
404
Sample Matrix
Condensate
Condensate
Condensate
Condensate
Condensate
Condensate
Condensate
Condensate
Transformer Oil
Transformer Oil
Transformer Oil
Transformer Oil
Transformer Oil
Transformer Oil
Transformer Oil
Coolant
2,4-D Rinse Water
Waste Solvent
Herbicide
Paint/Solvent
Waste Solvent
Waste Solvent
Waste Oil
Chlor. Solvent
Paint
Pump Oil
Waste Solvent
Herbicide
Paint/Solvent
Printing Solvent
GC Results
Aroclor
NDb
ND
1242
1242
1242
1242
1254
1242
1260
PCBC
PCB
PCB
PCB
PCB
PCB
PCB
1254
1242
ND
1254
1242/1260
1242/1260
1260
ND
ND
ND
ND
ND
ND
ND
Cone, ppm
ND
ND
25
5
<10
58
25
200
183
20
38
163
176
336
6400
10
<10
29
<2
9
11/17
2/2
323
<5
<50
<50
<35
<50
<5
<5
IA Results
Test
Results
<5
<5
>5
>5
<5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
<5
>5
<5
>5
>5
<5
>5
<5
<5
<5
<5
<5
<5
<5
Corr.
with GC
Results
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
                         4020-23
  Revision 0
January 1995

-------
                             Table 10 (continued)
Sample
ID
405
407
408
409
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
425
Sample Matrix
Waste Solvent
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
Waste Oil
GC Results
Aroclor
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
PCB
ND
ND
ND
ND
ND
ND
ND
ND
ND
Cone, ppm
<50
ND
ND
ND
ND
ND
ND
ND
ND
ND
50
ND
ND
ND
ND
ND
ND
ND
ND
ND
Number of False Positive Results
Rate
Number of False Negative Results
Rate
IA Results
Test
Results
<5
>5
<5
<5
<5
<5
<5
<5
<5
<5
>5
<5
<5
<5
<5
<5
<5
<5
<5
<5
Corr.
with GC
Results
yes
Fpd
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
1/32
3 . 1%
0/18
0 . 0%
a Trial  1  data

b ND =  Not Detectable

c PCB =  Aroclor was  not determined
d
  FP =  False positive
                                    4020-24
  Revision 0
January 1995

-------
                             Table 11
Correlation of PCB RISc™ Liquid Waste Test and Method 8082 Results

       Using Spiked and Unspiked Liquid Waste Field Samples
ID
001
002
003
004
005
006
007
008
009
010
Oil
012
013
014
015
016
017
018
019
020
021
022
023
024
025
026
027
Matrix
Aromatic solvent
Aviation gas
Chiller oil
Compressor oil
Coolant + water
Coolant oil
Coolant oil
Cutting oil
Cutting oil
Degreaser still
bottom
Dope oil
Draw Lube oil
Fleet crankcase
oil
Floor sealer
Fuel oil
Hi-BTU oil
Honing oil
Hydraulic oil
Hydraulic oil
Hydraulic oil
Machine oil
Mineral oil
Mineral spirits
Mineral spirits +
ink
Mixed flammables
Mixed solvents
Naphtha
GC
Results
Unspiked
ppm
<5
<5
<5
<5
<5
NRb
NR
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
NR
<5
<5
<5
<5
<5
<5
Immunoassay Result
Unspiked
ppm
<5
<5
<5
<5
<5
NR
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
>5
<5
<5
<5
Spiked (5
ppm 1248)
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
NR
>5
>5
>5
>5
>5
>5
Interp.























FP



                             4020-25
  Revision 0
January 1995

-------
Table 11 (continued)
ID
028
029
030
031
032
033
034
035
036
037
038
039
040
041
042
043
044
045
046
047
048
049
050
051
052
053
054
055
Matrix
Oil
Oil
Oil
Oil
Oil
Oil
Oil + 1,1,1-
trichloroethane
Oil sludge
Oil + freon
Oil + mineral
spirits
Oil + scum
solution
Oily water
Paint thinner
Paint thinner
Paint thinner
Paint waste
Paint waste +
thinner
Perce + oil
Petroleum
distillates
Petroleum naphtha
Pumping oil
RAC-1 SKOS
Sk oil
Sk oil
Smog Hog
Toluene + hexane
Toluene + stain
1,1,1-
Trichloroethane
GC
Results
Unspiked
ppm
n <5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
NR
<5
<5
<5
<5
<5
Immunoassay Result
Unspiked
ppm
<5
<5
<5
<5
<5
<5
<5
>5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
>5
<5
<5
<5
<5
<5
<5
<5
<5
>5
Spiked (5
ppm 1248)
^ >5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
Interp.







FP










FP








FP
       4020-26
  Revision 0
January 1995

-------
Table 11 (continued)
ID
056
057
058
059
060
061
062
063
064
065
066
067
068
069
070
071
072
073
074
075
076
077
078
079
080
081
082
Matrix
1,1,1-
Trichloroethane
1,1,1-
Trichloroethane
1,1,1-
Trichloroethane
1,1,1-TCE +
methanol
Trichloroethylene
Trichloroethylene
Trichloroethylene
Turpentine
Used n-
butyl acetate
Used oil + freon
Used oil + freon
Used oils
Used petroleum
Used petroleum
Used synthetic oil
Varnish + stain
Varsol
Waste coolant +
oil
Waste ink +
solvent
Waste naphtha
Waste oil
Waste oil
Waste oil
Waste oil
Waste oil
Waste oil
Waste oil
GC
Results
Unspiked
ppm
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
Immunoassay Result
Unspiked
ppm
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
Spiked (5
ppm 1248)
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
Interp.



























      4020-27
  Revision 0
January 1995

-------
                             Table 11 (continued)
ID
083
084
085
086
087
088
089
090
091
092
093
094
095
096
097
098
099
100
Matrix
Waste oil
Waste oil
Waste oil +
kerosene
Waste oil + gas
Waste paint
Waste paint
Waste paint
Waste paint
Waste paint
Waste paint
Waste SC-49
solvent
Waste solvent
Waste stoddard
Waste toner
Waste tramp oil
Waste transmission
fluid
Xylene
Not Recorded
No. of False Positive
Results
Rate
No. of False Negative
Results
Rate
GC
Results
Unspiked
ppm
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
Immunoassay Result
Unspiked
ppm
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
<5
>5
<5
6/99
6.1%

Spiked (5
ppm 1248)
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
>5
NR
Interp.









FP






FP


0.98
0 . 0%
a Trial  2  data

b NR =  not run
                                    4020-28
  Revision 0
January 1995

-------
                                  METHOD  4030

           SOIL SCREENING FOR PETROLEUM HYDROCARBONS BY IMMUNOASSAY
1.0   SCOPE AND APPLICATION

      1.1   Method 4030 is a procedure for  screening soils to determine whether
total petroleum hydrocarbons (TPH) are likely to be present.  Depending on the
testing product selected,  samples may be used to locate samples with low (<40-100
ppm),  medium,   and  high  (>1000  ppm)  concentrations  of  contaminates,  or  to
determine if TPH  is present  at concentrations  above  5,  25,  100,  or 500 mg/kg.
Method 4030  provides  an estimate  for the  concentration of  TPH  by  comparison
against standards, and can be used to produce multiple results within an hour of
sampling.

      1.2   Using the  test kit  from which  this method was  developed,  95  % of
samples containing 25 ppm  or less of TPH will produce a negative result in the
100 ppm test configuration.

      1.3   The sensitivity of any immunoassay test depends on the  binding of the
target analyte to the antibodies used in the kit.   The testing product used to
develop this method  is most sensitive to  the  small  aromatic compounds  (e.g.,
ethylbenzene, xylenes, and  naphthalene)  found  in fuels.   Refer to  the package
insert  of  the  testing   product  selected  for  specific  information  about
sensitivity.

      1.4   The  sensitivity  of  the  test is influenced  by  the  nature  of the
hydrocarbon contamination  and  any degradation processes operating  at  a site.
Although the action level of the test may vary from site to site, the test should
produce internally consistent results at a  particular site.

      1.5   In  cases  where a more  exact measurement of TPH  concentration  is
required, additional  techniques  (i.e., gas  chromatography Method  8015 or infra-
red spectroscopy Method 8440) should be used.

      1.6   This method  is restricted to  use  by  or  under  the  supervision  of
trained  analysts.    Each  analyst  must  demonstrate  the  ability to  generate
acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Test  kits  are  commercially  available   for  this   method.    The
manufacturer's directions  should be followed.

      2.2   In  general,  the method  is  performed  using  an  extract  of  a  soil
sample.   Filtered extracts may be stored cold,  in  the dark.   An aliquot  of the
extract  and  an  enzyme-TPH  conjugate  reagent  are  added to  immobilized  TPH
antibody.  The  enzyme-TPH conjugate  "competes" with hydrocarbons  present in the

                                    4030-1                           Revision 0
                                                                  January  1995

-------
sample for binding to  immobilized anti-TPH antibody.  The test  is interpreted by
comparing  the  response produced  by a  sample  to  the  response produced  by a
reference reaction.
3.0   INTERFERENCES

      3.1   Compounds that are chemically similar to petroleum hydrocarbons may
cause a positive test (false  positive) for TPH.  The data for  the lower limit of
detection of  these  compounds are provided  in  Tables  1A and IB.   Consult the
information  provided  by  the  manufacturer  of the  kit  used  for  additional
information regarding cross reactivity with other compounds.

      3.2   Storage  and  use  temperatures  may  modify the  method  performance.
Follow the manufacturer's directions for storage and use.

      3.3   Appropriate standards must be used (/.e., diesel standards for diesel
analysis, JP-4 for analysis of JP-4, etc.), or excessive  false negative or false
positive rates may result.


4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay test  kit: PETRO RISc Soil Test (EnSys,  Inc.), EnviroGard™
Petroleum Fuels  in Soil,  (Millipore,  Inc.),  or equivalent.   Each commercially
available test kit will supply or specify the apparatus  and materials necessary
for successful completion of the test.


5.0   REAGENTS

      5.1   Each  commercially  available test  kit  will  supply or  specify the
reagents necessary for successful completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory material to  this  chapter,  Organic Analytes,
Section 4.1.

      6.2   Soil samples may  be contaminated, and should  therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions  for the  test kit  being used.
Those test kits  used must meet or  exceed the performance specifications indicated
in Tables 2-12.
                                    4030-2                          Revision 0
                                                                  January 1995

-------
      7.2   Appropriate  standards  must  be used to  prevent  excessive rates of
false negative or false positive results.


8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit  being used
for quality control  procedures specific to the test  kit  used.   Additionally,
guidance provided in Method 4000 and Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly  when  results indicate
concentrations  near  the  action  level,   is  recommended to  refine   information
gathered with the kit.

      8.3   Do not use test kits past their expiration date.

      8.4   Do not use tubes or reagents  designated for use with other test kits.

      8.5   Use the  test kits  within their specified  storage temperature and
operating temperature limits.

      8.6   Method 4030  is intended for field or laboratory use. The  appropriate
level of quality  assurance  should  accompany the  application of this method to
document data quality.


9.0   METHOD PERFORMANCE

      9.1  A single laboratory study was  conducted with the PETRO RISc Soil Test,
EnSys, Inc., using five contaminated soil samples.  The samples were contaminated
with oxygenated gasoline, oxygenated gasoline  24 hours after  contamination, low
aromatic diesel (purchased in California),  normal diesel (purchased  in Northern
Virginia),  and JP-4 jet  fuel.  Five replicate determinations  were made using the
kits, and the data compared  with  values  obtained using GC/FID (Method 8015) and
IR  (Method  8440).   Several  different  analysts ran  the  immunoassay analyses.
Samples two- to five-fold below  the  action level  generally  gave readings less
than the action level.   Samples  two  fold above the  action level  gave readings
greater than the  action level.   Samples at or near the action level  give mixed
results (e.g., both less than and greater than  the action  level).  Tables 2 - 6
summarize these results.

      9.2   Sensitivity of the  EnviroGard Petroleum  Fuels  in  Soil Test Kit  was
determined  by  establishing  the   "noise"  level  expected  from matrix  effects
encountered in  negative  soil  samples   and  determining  the  corresponding  TPH
concentration by comparison  to the analyte-specific response  curve.   8 different
soils which did not contain  TPH were assayed.  Each of these  soils was extracted
in triplicate and each extract  was assayed  in three  different assays.  The mean
and  the  standard  deviation of  the resulting  %Bo's (%Bo  =  [(ODsample/ODnegatlve
control)xlOO]) were  calculated and  the  sensitivity was estimated at  two standard
deviations below the  mean.  The sensitivity for Method 4030 was determined to be


                                    4030-3                           Revision 0
                                                                  January 1995

-------
80% Bo at a 95% confidence interval.  Based on the average assay response to home
heating oil (HHO), this corresponds to 5.8 ppm.  These data are shown in Table
7.

      9.3   The effect of water  content  of  the soil  samples was determined by
assaying three different soil  samples which  had been dried and  subsequently had
water added to 30%  (w/w).   Aliquots  of these  samples  were then fortified with
HHO. Each soil  sample was assayed three times,  with and without added water, and
with and without HHO fortification.  It was  determined that  water  in soil up to
30% had no detectable effect on the method.   These data are shown in Table 8.

      9.4   The effect of the pH of the soil  extract was determined by adjusting
the soil pH of three soil samples.  Soil samples were  adjusted to pH 2 - 4 using
6N HC1 and pH 10 - 12 using 6N NaOH.  Aliquots of the pH adjusted soil  samples
were fortified with home heating oil.  Each soil sample was assayed unadjusted
and with  pH  adjusted to 2-4 and 10-12,  both  unfortified  and fortified.  These
extracts were assayed three times.   It was determined  that soil  samples with pH
ranging from 2 to 12 had no  detectable effect  on the performance of the method.
These data are shown  in Table 9.

      9.5  Two field studies were conducted at contaminated sites using the PETRO
RISc Soil Test, EnSys,  Inc..   In  Field Trial  1,  the  method  was used to locate
soil contamination resulting from a leaking above ground gasoline tank.   In Field
Trial 2, the  method  was used to evaluate diesel  fuel contamination  in a railroad
contaminated soil,  sludge,  and wastewater impound.   Overall,  a high degree of
correlation was observed between the standard method and the  immunoassay method.
The application of the immunoassay method to 23 samples (46 analyses) resulted
in  eight  conclusive  false  positive  results  (17%) and  three  conclusive  false
negative  results  (7%).   Tables 10 and 11 summarize these results.   There was
agreement for 76% of the samples tested in the two trials  for which  data are
presented.

      9.6   Two field trials were undertaken to investigate the ability of the
EnviroGard Petroleum Fuels  in  Soil  Test Kit  to  identify soil samples which were
contaminated with TPH.  In trial 1 the method was used  to identify soil which was
contaminated  with  gasoline from  leaking  underground  storage  tanks.    The
immunoassay was compared to Method 8015.  Twenty samples were analyzed by both
methods.  Interpreting the  results at a cutoff of  100  ppm resulted in 1/20 (5%)
false negatives and 0/20 (0%)  false positives.   In  trial 2,  the  method was used
to  identify  soil  which was  contaminated  with  JP-4 jet  fuel  from  leaking
semi-submerged storage tanks.   The  immunoassay was compared to Method 8015.  Ten
samples were analyzed by both  methods.   Interpreting  the  results at 1,000 ppm
resulted  in 0/10 (0%) false negatives and 1/10  (10%)   false positives.  Overall,
for both  field trials,  there were  1/30 (3.3%)  false  negatives and 1/30 (3.3%)
false positives.  These data are summarized in Table   12.


10.0 REFERENCES

1.    PETRO RISc™ Users Guide, Ensys Inc.


                                    4030-4                          Revision 0
                                                                  January 1995

-------
2.    Marsden, P.J.,  S-F Tsang, and N. Chau,  "Evaluation  of the PETRO RISc™ kit
      Immunoassay  Screen   Test  System",  Science  Applications  International
      Corporation under contract to EnSys Inc.,  June 1992,  unpublished

3.    EnviroGard™ Petroleum Fuels in Soil Test  Kit  Guide, Millipore,  Inc.
                                    4030-5                           Revision 0
                                                                  January  1995

-------
Table 1A
CROSS REACTIVITY3
Compound
Gasol ine
Diesel fuel, regular #2
Jet A fuel
Kerosene
Fuel oil n
Mineral Spirits
Light lubricating oil
Lithium grease
Brake fluid
Chain lubricant
Toluene
o-Xylene
m-Xylene
p-Xylene
Ethyl benzene
Hexachlorobenzene
Trichloroethylene
Acenaphthene
Naphthalene
Creosote
2-Methylpentane
Hexanes, mixed
Heptane
iso-Octane
Undecane
Soil Equivalent Concentration (ppm)
Required to Yield a Positive Result
100
75
75
100
100
<30
7,000
10,000
>10,000
>10,000
200
50
100
300
50
<30
1,000
<30
<30
<30
150
250
300
30
>10,000
PETRO RISc Soil Test, EnSys, Inc.
                             4030-6
  Revision 0
January 1995

-------
TABLE IB
CROSS REACTIVITY3
Compound
1,2,4 - Trimethyl benzene
m - Xylene
Acenaphthylene
Acenapthene
p - Xylene
Naphthalene
1,3,5 - Trimethyl benzene
Fl uorene
Phenanthrene
o - Xylene
Ethyl benzene
Toluene
Propyl benzene
Chlordane
Benzene
Toxaphene
Concentration Required
Positive Interpretation
for
(ppm)
0.1
0.3
0.3
0.4
0.5
0.7
2
2
2
3
5
7
11
45
70
70
The following compounds were tested and found to yield negative results
for concentrations up to 1000 ppm:
PCB (Aroclor 1248) TNT
Pentachlorophenol DDT
EnviroGard™ Petroleum Fuels in Soil, Millipore,  Inc.
                                 4030-7
  Revision 0
January 1995

-------










^3-
i
Q_
*™D
g«
£-
S*
=3 •£
00 S
UJ U
o: ^
j— •.
1 g-
GJ
CM S-
UjLO
[
CO

05 £Z
£= CD
•r- (J
E C
0 0
Z 
CD
s:
E
CL
Q.
0
o
r~-
CM
t
0
r-..
ro
<-— ^^^
Q. S-
CL CD
O r—
CM +•>
IO 3
i 0
o
00 r-H
ro -^









E
CL
CL
LO
en
1
CM
LO






CL
CL
ro
LO
00
CM











«=C













•o
C£
t— t
>,
^Q
•o
CD
C
i
CD
+->
CD
•o
CD
S-
CD
S
00
c
o
•r—
-t->
ro
oncentr
o

r—
ro
c
•r—
E
O
"Z.


CD
00
rO
GO
CD
O
S-
o
u.

i-
•r-


ro

-»->
ro

00
CD
s-
o
o
oo
ro
c
CD
^.
ro
+->
CD •
S- oo
CD T-
S oo
>!
00 1 	
QJ ro
r— C
CL ro
E
ro Q
OO i— i
Ll_
^^
• O
03 C3
-4->
Jx£
CD
JZ
+->
-C
+-1
S
-o
CD
CL
CL
3
oo
c:
0
-O
s-
ro
O
O
s_
•o
>5
-C
CD
-C
-(->
0)
£1
00
3
0
ro
O
^~
•u
o
_£Z
-(->
CD
^
o
+->
O)
c
., —
-o
s_
o
u
u
ro
c
3
S-

ro
3
+->
oo
CD
+->
5
r-
u|
CO
1— 1
o;

O
ce:
I —
UJ
CL.


.
_Q


•o
CD
N
>>

'(fl
C
rO

+->
O
d

00
ro
S
-t->
u
ro
S-
-l->
X
CD

E
CL
CL
O
CM
T3
tandar
00

^~
1
Q.
O

ro
CD
C

00
^
c:
3
s_
00
ro
^

LO
r_(
O
00

TO
o
_c
+- >
CD
2:


.
(_>
o
>)
+->
_d
ro
S-
ro
>
CD •
-C CD
+-> i —
CL
<+- E
O ro
00
CD
oo E
3 CL
rO Q.
U
CD CD
CO CM
CD
• J=
T3 +->
O
-C C
+-> O
CD
E -0
CD
CD +->
-C O
4-> 3
-a
IZ C
-r- O
o
T3
CD QJ
•r- i_
4- CD
•r- 2
0
CD 00
CL CD
00 00
>,
00 i —
•*-> ro
c c:
CD ro
>
r— O
O "Z.
^r- •
0 T3
CD
CD -t->
s_ s_
3 O
4-> Q.
X CD
•r- S_
E
00
CD -r-
-C
+-> 00
CD
CT> 3
C i —
•i— ro
00 >
3
f[
C O
^
i- CD
Dl
00 C
ro ra
S S-
.—i CD
• JZ
00 -t->
r-H
^H^ »^
OO
"O -t->
O r—
<— 3
-r^> OO
QJ QJ
2: s-

QJ
• JT
TD +->
 O  LO
     en
 C  CTl
 o  ^

 00  >•>

 >  ro
 QJ  3
 cc:  c:
     ro
oo

o
ro
O

-------

1
LU
OO
UJ
t— t
0
<_J
1— t
1—
<
0 +->
QC 00

3 \
O oo
—1 QJ
-t->
CC ro
O 0
U_ •!—

OO CL
1 — QJ
— 1 S_
^
oo in
LU
o;
!

^^
C*)
LU
— 1
co
eC
| —
























g
Q_
Q.
O
O
O


E
CL
CL
O
in




E
CL
CL
m
o
i— i




g
0
CL
in
r —




e
CL
CL
in
CM
r— 1




(0
C
O
•^~
+->
ro
S-
i — J->
ro C
C QJ
•r- U
E c
0 0
2: o
o
o
UD
I-H
A
in E
\ CL
in CL


E
CL
Q.
~o
in in
\^ r-H
ro A


E
CL
CL
-0
m in
^s. f-~<
in v



o
in
i — i
v

*3- E
\ CL
•3- CL



£^
in

v
r»
^1- E
\ CL
•cl- CL




f)
2
\—
U|
OO
t— 1
a:
o
DC
1—
LU
Q_
m
O
+ 1
0 E
UD CL
CD CL


CM
1 — 1
+1
in E
CM CL
•—i CL
E
CL
Q.

m
i— i
+1
0
CT>


£
CL
CL
r--.

+ 1
*3-
m








-a
c





o
in

1— H
O
oo
-o
o
-C
•!->
QJ
2:


0
O •
• 00
O CD E
t— i 1 — CL
00 •— < CL


 m E
CM O CL
•— i CO CL



E
CL
i CL

0 0
l£> CM
O CT)
r-H CM


r-^

, — 1
in
'
in
• E
O CL
CO CL








T>
a;
t-H
C
•1 —
-o
'o
00
OJ
00
QJ
•r-
-a
o
4->
ro
E
o
S~-
ro
-3.
O
J —
U—
0
00
-4-J
£Z
3
O
E
rO
C
3
o
_^
^1^
+j
3

, —
•r—
O
00
E
ra
o

>,
"O
c:
ra
00

CD
C
_*:
•i —
CL
00 -~~
Qi
>l^-J
-Q 0

-o <-
QJ ID
i- in
ra CM
CLCM
QJ
4_ c
CL o
,r—
QJ +J
S- O
QJ QJ
3 oo
C/>
OJ n3
CX C
E ^~
rrs O
CO 4-
.,
i —
• fO
n4 t ^
in
i— i
*•>
-t->
^i
a>
jc
-»->
JC
4->
3
-o
QJ
•r—
'a.
CL
3
00

rocarbon
-o
>>
jc
QJ
-C
+J
CD
C
oo
3
O
ro
O
^j-
-0 vi
0 -i-1
51
QJ -r-
!g" i —

O <->
+3 0-
CDH-
c o

^3
0 0
0
0 E
ro CL
CL
c
3 m
S- r-
oo "O
ro C
3 ">
4j in
oo ro
QJ
+-> -t-1
ra
S
K oo
01 C
oo o
t— 1 '^-
Di +->
ra
o c
C£ -r-
1— E
LU S-
f\ QJ

•4^
* QJ
o -i— i
tion. nd - not detected.
ro
-I-)
00
ra
•p —
c:
i-
O
<4-
"'
rO
O
rO
+->
ro
-o
QJ
00
ro
.C
o
S-
3
CL
-o
ro
"O
C
rO
-,->
00
i
QJ
00
OJ
•1—
-a

ra

CT)
c:
•i —
oo
3
C
3
S-
00
ra
3

in
o
oo
•o
0
_c
+->
QJ
^"


U
JO
ra
•r- QJ
S- i —
ra CL
> E
ra
QJ 00
-C
•M E
CL
4- CL
O
O
QJ m
oo i— i
3
rO QJ
0 -C
QJ +->
CO
S-
0
• M-
-o
O QJ
-C T3
-I-J ra
QJ E
C
QJ O
.C T-
+-> +->
ro
C C
g
-o s_
QJ OJ
•r- -(J
14- a>
•r- -O
o
QJ ac
CL>-H
00
o
00 C
•4->
C 1
QJ
> 
QJ S-
S- O
3 CL
-t-> QJ
X S-
•r~
E oo
QJ
-C 00
•+-> -4-J
i 	
CD 3
c: oo
•r- OJ
00 S_
3
*4—
C O
3
S- QJ
CT)
oo c:
ra ro
3 S-

i — I QJ
• -C
CO +J
i — i
•*r ~
00
-o -M
O r—
JZ 3
-I-J 00
QJ QJ
2: s_
nj

T3 -r->
o un
   en
 c= en
 O -—i

 00  >,

 >  ro
 QJ  3
a:  c
    ro
 en

 O
 ro
 O

-------
UJ
oo
LU
Q
o;

p— H



E
CL
CL
in
r~-







E
CL
CL
in
C\J



ra
c
o
+->
m
i.
i — +->
rO <~
C QJ
•r- (J
E C
0 0
Z 0


E
CL
CL

in
r —
A
•^r
^-

E
CL
CL
in
i — *
A
^
ro
C\J




E
CL
CL
in
t^
V
*%
_^
^^
\^
CM




&
5
i—
0|
oo
h— 4
C£
O
a:
i—
LU
Q.


E
CL
CL

*S-

O
i — i
+ 1
CVJ
10
l~*

E
CL
CL
00
r-.
+ |
CT»
in
i~-


^
CL
CL
*3-
*&

+ 1
csj
r-H
in




o
in
i — i
o
00

"O
o
_c
4->
OJ
s:
OJ
oo
OJ
T3
CM
S-
OJ
-Q
E
3
C
L-
CO
3
CD
OJ
<+-
O
oo
4-*
C
3
O
E
rC

c
s
o
c
.^
_c
-t->
•i —
2
O
00
E
ro
O
r~
>i
T^
C
ro
00

CD
c:
-l~
-^
CL
to

>>
-£2
"O
QJ
S-
ro
CL
QJ
S-
CL
O)
i_
QJ
5
00
Q)
CL
E
ro
OO

.
QJ
C
O
+->
-^
QJ
.C
4->
^:
+->
*t—
2
•o
QJ
CL
Q.
3
00
E
O
.0
S-
rO
(_)
O
^_
-o
>>
^

OJ
_^
1 1

en
c
00
ZJ

CD
00
O
•^J-
"O
o
jc oo
4-> +->
OJ '^
s: E
o i—
 CL
00
QJ m
-t-> i~-
5 §
0|
00 C
1—1 O
Q; -.-
4-1
O ro
Q; c
LU i
Q- S_
QJ
+J
• QJ
a station.
c
CD
S^

>

ro

-t-J
ro
T3
QJ
00
ro
-C
O
s-
3
CL
-D
S-
ro
"O
sz
ro
j 1
00
, —
QJ
00
QJ
•i—
"O
ro
CD
sz
00
ZJ
£Z
Z!
S-

00
ro
3
in
o
CO
T3
O
-C
+->
QJ
^

.
                                                                                                o un
                                                                                                   CTl
                                                                                                c: cri
                                                                                                O f—i

                                                                                                oo  >,


                                                                                                >  ro
                                                                                                QJ  3
                                                                                               CtL  C
                                                                                                    ro
                                                                                              CD
                                                                                              n
                                                                                              o
ro
                   O

-------




















LU
^
t— (
Q.
OO

^
OO
LU
dL
u_

i

LU
•z.
*~~*
	 1
O
oo
et
CJ3

Q
LU
1—
-
X
o

CSL
O
u_

OO
1 —
— 1
^
00
LU
C£

I

LO

LU
	 1
CO

(SI
QJ
4->
C/>
QJ
+J
to
o
•r—
'a.
O)
S-
LO





















































E
CL
CL

O
0
o
r— <













E
CL
CL

0
O
OJ















E
CL
CL

O
O


















E
CL
CL

O
LO
















, —
ro
C

E
O
Z.





















































































CO
C
O

+->
to
S-
-t->
C
QJ
O
c
o
u
E
CL
CL

CD
CD
CD
i— (
A

t*
LO
\^
LO







E
Q.
CL

CD
CD

A

r-
LO
^*^_
LO









E
CL
CL

O
0
i— i
A
^
LO
^3-










E
CL
CL

CD
CD
i — i
V

r.
"3-

CO










-Q
5
H
<->|
oo
i — i
o:

o
0£
i—
LU
CL.


E
CL
CL

IO
OJ

+1

**•
CO
**•





E
CL
CL

cn
•
O
i— i

+i

00
•

QJ
SI
QJ
S_
O)
3

QJ

CL
E
rO
is>

*.
, —
O)
Z3
l+-
-o
0)
+^
ro
c
QJ
0>
>>
X
0

c.
r«

4-
O

(/)
-(->
C
13
O
E
to

c

o
c
.^

_c
4->
•r—
•5

o
I/)
E
to
o


>-,
-o
c
rO
(Si

CD
c
• 1 —
_*:
.,^
CL
(>0

>,
-Q

-o
O)
s_
to
Q-
Q)
S-
CL

OJ
S_
o>
3

iy>
QJ

CL
E
rO
00



ra


























































.
-o
ai
N
>>

ro
c
rO

J—
-1 —
+->
c
3

i/i
s_
to
•i-J

-o
OJ
(SI
O
1^
o

c:
• r—

T3
QJ
C
• i —
ro
4->
C
•1 —
ro
E



QJ
c
O

„
1 1
•r-
J*.

QJ
_cr
+j
^:
4->
3
-^
O)

, 	
CL
Q.
3
(SI

c
O
_Q
s_
ro
<_)
O
s_
T3
>,
r-

QJ
f~
4->

CD
C

(/)
13

o
CO
O
•sr

-o
o

-t_)
QJ
2;

o
1 ^

CD
C
-i —
-o
s_
o
u
(_J
rO

c:
^
S-

i/i
rO
3

-t-J
iyi
QJ
4->

5
1—
01
OO
t— 4
QL

O
Di
I—
LU
Q_



-Q




















































I/)
-t->

E

,^

O
CD"

<+-
O

-t->
3
0

QJ

'Q.
E
to
c/1

E
CL
CL

CD
LO

C
o

c:
o
•i —
+->
to
sz
•( —
E
S-
QJ
•+->
OJ
T3




















C
o
•r—
+J
ro
+->
(SI

rO
• f—
c:
s-
o
q-
•i —
i —
ro
0

rO

4->
to
-o
QJ
(SI
ro
-C
(J
s-
3
CL

-o
S_
rO
-o
c:
to
+->
(SI

O)
c
•l~
r—
O
00
to
CD

rO

CD
C
• 1 —
(Si
13

C
rj
s-

(xi
to
3

LO
i — 1
O
CO

TD
O
c~
4-J
QJ
21



O
O  LO
    CTl
 c  en
 o  -—i
•^  ^
 >  ra
 OJ  3
cc:  c
    rO
O
ro
O

-------
                                                              0)







E
CL
CL
O
O
Cvl




E
CL
CL

O
o
r— H
A

•"> cH
r— ^
<° ~S
c £
ro g
U
i^
^ n«
(^ CU
-^
-o -^
<" n
IT)
^^
1-1
^
+->

_i^

— i CL
° S
oo _ o
 E •— '
cj co CL A
O CL
Q J_) "
LU ^ 0 ID
| — oo CD \
ef CO 1~H ^~
"Z. +->
LU ro
C3 O
>- -r-
X i—
0 CL
CO
cn s-
o
I_L_ 1_O
00
1 —
-=>
00 p.
LU fc-
ct °-
"^ CL
' O
^ 2
E V
m °- /-
^ o \
LT) OO


>>
X
o
c
ro
Ot-
E 0
CL
CL oo
+->
c
•*!- 3
O
0 E
rC
+1
10 3
o
oo a
.^

_^
4->
•i —
3
i —
o
00
E
ro
O
>,
T3
c
ro
00
cn
c

-^£
-o -^
C Q.
1/1

LJ.
O
•o
CO
s_
0
-i->
00
CO
s_
CO
s
1/1
CU
r—
CL
E
rO
oo

^j
ni
tu
\s

'1~~
ri
OO

.
CO
'Q-
ro
00
CO
_c:
-t->
cn
c:
tsl
••"
d
OJ
cn
o
E
r~>
















O)
>
CO
r^
cn
£=
sx
CL
00
CO
-C
-+— >

hydrocarbon supp

OJ
.c
+->

cn
c
«i—
00
=3

O
CO
o
•"• J>
-o -^
5i
4-> '•"
O) •"
^0
o °r
•^M-
cn 0
= +J
1 o
0 !=
y i-
re °-

c °
•r—
+->
rO
-t-)
00
ro
C
S-
o
14-
ro
t_>

ro
+->
rO

-o
CO
oo
ro
r~
U
S-
3
CL
T3
S-
ro
-a
c:
ro
-fj
oo
CU
C
•i —
O
00
ra
cn
ro

cn
                                                                                                                                    o LO
                                                                                                                                       cr>
                                                                                                                                    c cn
                                                                                                                                    O r-H

                                                                                                                                    00  >,

                                                                                                                                    >  ro
                                                                                                                                    CU  ^
                                                                                                                                    cn  c
                                                                                                                                        ro
                                                                                                                                   CD
                                                                                                                                   ro
                                                                                                                                   CD
           c
           o
 ro
 S-
4->
 C
 a>
 c_i
 £Z
 O
 o
                   oo     o
                   >-i     CO
           E     r-
           O     LU
                          "O
                           O
                           a>
                                   a,  co  co
                                   c. •*->  oo
ro
CL
co
S-
                                       oo  O)
                                       S-  S-
                                       3  ro
                                   O>
                                      ^t-  o
                                      CNJ -^
                                      "O  ro
  "  O  C
    3  CU
   T3  O
    c:  c
    o  o
    (J  U
               00
                                                      ai
                           00
                           ra
                           3
                                                       CU
                                                       00
                                                                         a>
                                                                         o
                                                                         ai
                                                                         QJ

-------
                                      TABLE 7
                                METHOD  SENSITIVITY

                   Part 1  - Average Response with Negative Soils

Soil#       Soil Type          Average %Bo (n = 9)Standard Deviation
SAND
S2
S3
S4
S5
S6
S7
S8
91.4
LOAM
CLAY
LOAM
CLAY
LOAM/SAND
SAND/LOAM
LOAM
4.1
83.1
84.4
80.9
89.7
91.2
89.0
90.0
3.2
3.1
1.3
1.7
0.2
0.3
1.4
AVERAGE                            87.5                 4.0


                    Part 2 - Average Response with  Calibrators

    Calibrator
    Concentration  (ppm)    Average Absorbance           Average %Bo

0
5
15
50
125
1.339
1.097
0.825
0.427
0.219
                                                          N/A
                                                           81.9
                                                           61.7
                                                           31.9
                                                           16.3
                            Part 3 - Method Sensitivity

Based on Part 1 and Part 2  Above:
         Average %Bo - 2 SO =  79.6  which  is  equivalent to 5.8 ppm
         Average %Bo - 3 SD =  75.6  which  is  equivalent to 7.0 ppm

(%Bo = [(ODsample/ODnegatlvecontro,)xlOO])
                                    4030-13                          Revision 0
                                                                   January 1995

-------
                                       TABLE 8
                      EFFECT  OF  WATER CONTENT IN SOIL SAMPLES
Soil % Water Fortified?  Rep.  1* Rep.  2  Rep. 3  Mean Std. Dev.  ± 2 SD  Range
SI
SI
SI
SI
S2
S2
S2
S2
S3
S3
S3
S3
0
30
0
30
0
30
0
30
0
30
0
30
No
No
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
101
100
59
60
57
74
40
44
70
81
41
61
.3
.5
.2

.9
.5
.3
.5
.1
.5
.1
.3
99
115
65
74
53
91
40
67
85
109
46
76
.1
.5
.8
.7
.9
.8
.9
.8
.6
.4
.6
.7
111.8
109.1
69.6
83.1
72.3
85.2
45.6
68.4
76.7
103.4
60.7
63.1
104.1
108.4
64.9
72.3
61.4
83.8
42.3
60.2
77.5
98.1
49.5
67.0
6.8
7.5
5.3
11.7
9.7
8.7
2.9
13.6
7.8
14.7
10.1
8.4
90.4
93.4
49.9
49.2
42.0
66.4
36.5
33.0
61.9
68.7
29.3
50.2
- 117.7
- 123.4
- 75.5
- 96.0
- 80.8
- 101.2
-48.1
- 87.4
- 93.1
- 127.5
- 69.7
-83.8
* All values  shown are %Bo  [=  (ODsample/ODn
                                          egative coi
,n«roi)xlOO]
                                     4030-14
                       Revision 0
                     January 1995

-------
                                        TABLE 9
                             EFFECT OF pH  ON  SOIL SAMPLES
oil
SI
SI
SI
SI
SI
SI
S2
S2
S2
S2
S2
S2
S3
S3
S3
S3
S3
S3
oH Ad.i. Fortified? Rep.
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
88.
108.
101.
64.
52.
69.
76.
101.
89.
59.
68.
47.
83.
89.
80.
60.
58.
53.
1*
9
9
2
3
9
3
2
2
9
4
1
8
4
3
6
2
8
4
Rep. 2
93.2
66.0
90.3
55.7
41.1
61.7
86.4
82.4
72.1
60.3
62.3
51.7
88.4
84.9
84.2
53.6
58.5
41.8
Rep. 3
92.8
88.1
90.6
58.0
49.4
57.5
83.1
99.5
77.7
53.7
59.3
39.4
85.3
91.0
90.3
58.8
62.0
59.9
Mean Std. Dev.
91.6
87.7
94.0
59.3
47.8
62.8
81.9
94.4
79.9
57.8
63.2
46.3
85.7
88.4
85.0
57.5
59.8
51.7
2.4
21.5
6.2
4.5
6.1
6.0
5.2
10.4
9.1
3.6
4.5
6.3
2.5
3.1
4.9
3.5
1.9
9.2
± 2 SD
86
44
81
50
35
50
71
73
61
50
54
33
80
82
75
47
56
33
.8 -
.7-
.6 -
.3 -
.6 -
.8 -
.5 -
.6 -
.7 -
.6 -
.2 -
.7 -
.7 -
.2 -
.2 -
.7 -
.0 -
.3 -
Range
96.4
109.2
106.4
68.3
60.0
74.8
92.3
115.2
98.1
65.0
72.2
58.9
90.7
94.6
94.8
64.5
63.6
70.1
* All values shown are %Bo  [= (ODsample/ODn
                                           egative co
n,rol)XlOO]
                                      4030-15
                       Revision 0
                     January 1995

-------
Table 10
PETRO RISc Soil Test
Field Trial 1
Sample ID
AST-01
AST-02
AST-03
AST-04
AST-05
AST-06
AST-07
AST-08
AST-09
IR Method (ppm)
<20
520
1700
130
20
40
400
640
1600
100 ppm Test
Result
< 100
> 100
> 100
> 100
> 100
> 100
> 100
> 100
> 100
AGREEMENT3
Y, FP, FN
Y
Y
Y
Y
FP
FP
Y
Y
Y
1000 ppm Test
Result
< 1000
> 1000
> 1000
< 1000
< 1000
< 1000
< 1000
< 1000
> 1000
AGREEMENT3
Y, FP, FN
Y
FP
Y
Y
Y
FN
FN
FN
Y
Y       Immunoassay  and  GC  or  IR  results  agree
FP      False  Positive
FN      False  Negative
                                  4030-16
  Revision 0
January 1995

-------
Table 11
PETRO RISc Soil Test
Field Trial 2
Sample
ID
1-B
2-A
2-B
2-C
3-B
3-C
4-A
4-B
5-A
5-B
5-C
6-B
8
9
GC
Extractables
(ppm)
5720
610
370
2270
4870
760
66
303
20400
26300
267
550
59300
26500
TRPH
(ppm)
20800
14700
6800
1950
18600
1180
4100
2100
29600
28600
330
22700
64400
12900
75 ppm Test
Result
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
> 75
AGREEMENT3
Y, FP, FN
TRPH
Y
Y
Y
Y
Y
Y
FP"
Y
Y
Y
Y
Y
Y
Y
GC
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
750 ppm Test
Result
> 750
> 750
> 750
> 750
> 750
< 750
< 750
< 750
> 750
> 750
> 750
> 750
> 750
> 750
AGREEMENT3
Y, FP, FN
TRPH
Y
FP
FP
Y
Y
FN'
Y
Y
Y
Y
FP
FP
Y
Y
GC
Y
Y
Y
Y
Y
FN
FN
FN
Y
Y
FP
Y
Y
Y
Y       Immunoassay  and  GC  or  IR  results  agree
FN      False  Negative
FP      False  Positive
FN*      False  Negative,  but within  25% of GC or  IR  results
FP"      False  Positive,  but within  25% of GC or  IR  results
                                  4030-17
  Revision 0
January 1995

-------
                     TABLE  12
        IMMUNOASSAY COMPARED TO METHOD 8015

Field Trial  1: Gasoline (Interpretation at 100 ppm)
Sample ID
MW-18-1
MW-18-2
MW-18-3
MW-18-A1
MW-18-A1 Duplicate
MW-18-A2
DBS
MW-12-3
MW-13-1
MW-13-3
MW-13-4
MW-17-3
MW-17-4
MW-17-5
MW-16-2
MW-16-2 Duplicate
MW-19-2
MW-19-3
MW-14-1
MW-17-A1
Method 8015 (ppm)
270
15
15
20
15
1500
300
250
40
50
20
250
180
180
11,500
11,500
10
70
280
560
Field Trial 2: JP-4 Jet Fuel
Sample ID
TB1 6.5-7.0
TB2 6.5-7.0
TB1 5.0-5.5
TB2 5.0-5.5
TBS 5.0-5.5
TBS 6.5-7.0
TB5 5.0-5.5
TB5 6.5-7.0
TB4 6.5-7.0
TB4 5.5-6.0
Method 8015 (ppm)
15,900
16,800
900
100
ND(<5)
29,500
5,000
2,000
19,000
5,900
Immunoassay
Negative
Negative
Negative
Negative
Negative
Positive
Positive
Positive
Negative
Negative
Negative
Posi ti ve
Positive
Positive
Positive
Positive
Negative
Negative
Positive
Positive
(Interpretation at
Immunoassay
Positive
Positive
Negative
Positive
Negative
Positive
Positive
Positive
Positive
Positive
Concurrence?
False Negative
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
YES
1,000 ppm)
Concurrence?
YES
YES
YES
False Positive
YES
YES
YES
YES
YES
YES
                      4030-18
  Revision 0
January 1995

-------
                                 METHOD  4035

      SOIL  SCREENING  FOR  POLYNUCLEAR AROMATIC HYDROCARBONS  BY  IMMUNOASSAY


1.0   SCOPE AND APPLICATION

      1.1  Method 4035 is a procedure for screening soils to  determine when total
polynuclear aromatic hydrocarbons (PAHs)  are present at concentrations above 1
mg/Kg.   Method 4035  provides an estimate  for the  concentration of  PAHs  by
comparison with a PAH standard.

      1.2  Using  the test kit from which  this method was  developed,  >95%  of
samples confirmed to  have concentrations of PAHs below  detection limits  will
produce a negative result in the 1 ppm test configuration.

      1.3  The  sensitivity of the test  is  influenced by the binding  of the target
analyte to the antibodies used in the kit.   The commercial  PAH  kit  used for
evaluation of  this method  is  most sensitive to the  three (i.e.,  phenanthrene,
anthracene, fluorene) and four (i.e. benzo(a)anthracene, chrysene, fluoranthene,
pyrene) ring  PAH compounds listed in Method 8310, and also recognizes most of the
five and six ring compounds listed.

      1.4  The  sensitivity of  the test is  influenced  by the nature of the PAH
contamination and any degradation processes operating at a site.   Although the
action level of the  test may vary from site to  site,  the  test should produce
internally consistent results at any given site.

      1.5    In  cases where  the  exact  concentration  of  PAHs are  required,
quantitative techniques  (i.e., Method 8310,  8270, or 8100)  should be used.

      1.6   This method  is  restricted to use  by or under  the supervision  of
trained  analysts.    Each  analyst  must  demonstrate  the  ability  to  generate
acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   An   accurately  weighed  sample is first  extracted  and  the extract
filtered using  a  commercially available  test  kit.  The  sample extract and  an
enzyme conjugate reagent  are added to immobilized antibody.  The enzyme conjugate
"competes" with the  PAHs  present  in the  sample for  binding to the immobilized
anti-PAH antibody.  The test is interpreted by comparing the response produced
by  testing  a  sample   to   the   response   produced   by  testing  standard(s)
simultaneously.

      2.2  A  portion  of all samples in each  analytical  batch should be confirmed
using quantitative techniques.


3.0   INTERFERENCES

      3.1  Chemically similar  compounds and compounds which  might be expected to
be found  in  conjunction with  PAH contamination were tested  to  determine the

                                   4035 - 1                         Revision 0
                                                                  January 1995

-------
concentration required to produce a positive  result.   These  data are shown in
Tables 1 and 2.

      3.2  The kit was optimized to respond to  three  and  four ring PAHs.   The
sensitivity of the  test  to  individual  PAHs is  highly  variable.   Naphthalene,
dibenzo(a,h)anthracene, and benzo(g,h,i)perylene have 0.5 percent or less than
the reactivity of phenanthrene with the enzyme conjugate.

      3.3  The  alkyl-substituted PAHs,  chlorinated aromatic compounds, and other
aromatic hydrocarbons,  such  as dibenzofuran, have been  demonstrated  to be cross-
reactive with the immobilized anti-PAH  antibody.  The presence  of these compounds
in the sample may contribute to false  positives.


4.0   APPARATUS AND MATERIALS

      4.1  PAH  RISc™ Soil Test (EnSys,  Inc.),  or equivalent.  Each commercially
available test  kit  will  supply or specify the  apparatus and materials necessary
for successful  completion of the test.


5.0   REAGENTS

      5.1   Each  commercially  available  test  kit  will  supply or  specify  the
reagents necessary for successful  completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material  to  this  chapter,  Organic Analytes,
Section 4.1.

      6.2  Soil samples may  be contaminated, and  should therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1  Method 4035 is intended for field or laboratory use.

      7.2  Follow the manufacturer's instructions for the test being  used.  Those
test kits used must meet or exceed the performance  indicated in Tables 3-7.

      7.3  The action  limit  for each  application must be  within  the operating
range of the kit used.


8.0   QUALITY CONTROL

      8.1  Follow the manufacturer's instructions for the test kit  being  used for
quality control procedures specific to  the test kit used. Additionally, guidance
provided in Chapter One should be followed.
                                   4035 - 2                         Revision 0
                                                                  January 1995

-------
      8.2   Use  of  replicate  analyses,  particularly  when  results  indicate
concentrations  near  the  action level,  is  recommended to  refine information
gathered with the kit.

      8.3  Do not use test kits past their expiration date.

      8.4  Do not use tubes or  reagents designated for use with other kits.

      8.5   Use  the  test  kits  within the  specified storage  temperature  and
operating temperature limits.


9.0   METHOD PERFORMANCE

      9.1   The extraction efficiency of a commercially available test kit was
tested (PAH RISc™ Test, EnSys Inc.) by spiking phenanthrene, benzo(a)anthracene
and benzo(a)pyrene  into PAH negative soil  matrices (PAH-116 and PAH-141  are field
samples).   The  soils  were spiked using  detection  limits  established  for each
compound (see  Table 1), extracted and determined by immunoassay.  The results for
these 3-, 4- and 5-ring PAHs (Table  4) demonstrated that they were extracted with
good recovery and yielded the correct assay interpretation.

      9.2  A single laboratory study was conducted with  a commercially available
test kit (PAH  RISc™ Test, EnSys  Inc.),  using 25 contaminated soil samples.  Four
replicate determinations  were made on each test sample and the data  compared with
values  obtained using  HPLC  Method  8310.    Several  analysts performed  the
immunoassay analyses.  The immunoassay data agreed in all cases with the external
HPLC data obtained (Table 5).

      9.3   An  additional  single laboratory validation  study  on  30  randomly
selected, PAH-contaminated field samples from multiple sites was run by  the USEPA
Region X Laboratory.   Results are reported in Table 6 on an  as  found basis, and
reported in Table 7 normalized  to phenanthrene,  based on cross-reactivity data
(from Table 1).  The false positive rate at the 1  ppm action level was 13% for
unnormalized results and 19% for normalized results based on 31 analyses.  The
false negative rate at 1 ppm was 0  in both cases.   At the 10 ppm action level,
the false positive rate was 19%  unnormalized and  26% normalized.  False negative
rates at 10 ppm were 6% unnormalized and 3% normalized.

      9.4   The probabilities of generating  false  positive  and false  negative
results at an action level of 1  ppm are listed  in  Table 3.


10.0  REFERENCES

1.     PAH-RISc™ Users Guide,  EnSys Inc.

2.     P. P.  McDonald,  R.  E. Almond,  J. P.  Mapes,  and S.  B.  Friedman, "PAH-RISc™
      Soil Test  -  A  Rapid,  On-Site  Screening  Test  for  Polynuclear  Aromatic
      Hydrocarbons  in  Soil",  J.  of  AOAC International (accepted  for publication
      document #92263)

3.     R. P. Swift,  J.  R. Leavell,   and  C.  W. Brandenburg,  "Evaluation  of  the
      EnSys PAH-RISc™ Test Kit", Proceedings, USEPA Ninth Annual Waste Testing
      and Quality Assurance Symposium, 1993.
                                   4035 - 3                         Revision 0
                                                                  January 1995

-------
Table 1: Cross-Reactivity of Method 8310 PAHs
Compound
2 Rinqs
Naphthalene
3 Rinqs
Acenaphthene
Acenaphthylene
Phenanthrene
Anthracene
Fluorene
4 Rinqs
Benzo(a)anthracene
Chrysene
Fluoranthene
Pyrene
5 Rinqs
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
D ibenzo( a, h) anthracene
6 Rinqs
Indeno(l,2,3-c,d)pyrene
Benzo(g,h, i)perylene
Concentration Giving
a Positive Result
(ppm Soil Equivalent)
200
8.1
7.5
1.0
0.81
1.5
1.6
1.2
1.4
3.5
4.6
9.4
8.3
>200
11
>200
Percent
Cross-Reactivity
0.5
12
13
100
123
67
64
84
73
29
22
11
12
<0.5
9.4
<0.5
                   4035
  Revision 0
January 1995

-------
   Table  2:  Cross  Reactivity of Other  PAHs  and  Related  Compounds
Compound
Other PAHs
1 -Methyl naphthylene
2-Methylnaphthylene
1-Chloronaphthylene
Halowax 1013
Halowax 1051
Dibenzofuran
Other Compounds
Benzene
Toluene
CCA
Phenol
Creosote
2,4,6-Trichlorobenzene
2,3,5,6-
Tetrachl orobenzene
Pentachl orobenzene
Pentachlorophenol
Bis(2-ethylhexyl)
phthalate
Aroclor 1254
Aroclor 1260
Concentration Giving
a Positive Result
(ppm, Soil Equivalent)
54
58
59
18
>200
14
>200
>200
>200
>200
5.4
>200
>200
>200
>200
>200
>200
>200

Percent
Cross-Reactivity
1.8
1.7
1.7
5.7
<0.5
7.2
<0.5
<0.5
<0.5
<0.5
18.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5

 Table  3:  Probability  of  False  Negative  and  False  Positive  Results
                 for PAHs at a 1 ppm Action Level
Spike Concentration
Phenanthrene (ppm)
0
0.4
0.8
1.0
Probability of False
Positive (Mean ± SD)
0% ± 0%
23% ± 17%
94% ± 13%
N/A
Probability of False
Negative (Mean ± SD)
N/A
N/A
N/A
0% + 0%
Results were obtained from spiking four different validation lots, using
3  operators,  12  matrices  for  a total  of 201  determinations  at  each
concentration of phenanthrene.

N/A = No false positive possible above action limit.
      No false negative possible below action limit.
                             4035 - 5
  Revision 0
January 1995

-------
Table 4: Spike Recovery of Phenanthrene, Benzo(a)anthracene and Benzo(a)pyrene
Compound
Blank
Blank
Phenanthrene
Phenanthrene
Phenanthrene
Phenanthrene
Phenanthrene
Phenanthrene
Benzo(a)anthracene
Benzo( a) anthracene
Benzo(a)anthracene
Benzo(a)anthracene
Benzo(a)pyrene
Benzo(a)pyrene
Benzo(a)pyrene
Spike
(ppm)
0
0
1
1
1
10
10
10
1.6
1.6
16
16
8.3
8.3
83
Soil
Wake
PAH-116
Wake
PAH-116
PAH-141
Wake
PAH-116
PAH-141
Wake
PAH-116
Wake
PAH-116
Wake
PAH-116
PAH-116
PAH RISC™
Results
<1
<1
1-10
1-10
1-10
>10
>10
>10
1-10
1-10
>10
>10
1-10
1-10
>10
                                   4035 - 6
  Revision 0
January 1995

-------
Table 5: Powerplant Field Samples (Soil) Evaluated by Immunoassay
Field Sample
Number
PAH-137
PAH-141
PAH- 118
PAH-136
PAH- 139
PAH-126
PAH-127
PAH-122
PAH-138
PAH-131
PAH-128
PAH-132
PAH- 112
PAH-140
PAH- 130
PAH-116
PAH-135
PAH-133
PAH-119
PAH-120
PAH-124
PAH-134
PAH-114
PAH-113
PAH-115
EnSys Method
Immunoassay (ppm)
>10
<1
1-10
>10
>10
1-10, >10
>10
>10
>10
>10
>10
>10
>10
>10
>10
<1
>10
>10
>10
>10
>10
>10
>10
>10
>10
Method 8310
HPLC (ppm)
<21
<21
<26
26
<28
<32
<33
<33
33
<34
<35
<43
<48
50
54
<61
71
<91
<100
<161
<167
182
<247
<294
<343
                             4035  -  7
  Revision 0
January 1995

-------
 Table 6:  Total  PAH Content of Region X Field Samples  Using  EnSys
                 PAH RISc™ Immunoassay Test Kit

Sample ID
PAH-1
PAH- 2
PAH -3
PAH -4
PAH- 5
PAH -6
PAH -7
PAH -8
PAH -9
PAH-10
PAH- 11
PAH- 12
PAH-12Dup
PAH- 13
PAH-14
PAH- 15
PAH- 16
PAH- 17
PAH- 18
PAH-19
PAH- 20
PAH-21
PAH- 2 2
PAH- 2 3
PAH-24
PAH-25
1 ppm Test
<1



*
*





*



*

*



*




*
>1
it




it


*
*





*

*
*
*



*
*

10 ppm Test
<10














*

*

*
*
*



*
*
>10
*
*
*


*
*
*
*
*

*
*
*



*



*
*
*



GC/MS
Lab Result
(ppm)'
0.2
12.2
16.0
0.0
0.5
8.7
148
182
4.4
0.2
0.0
85.4
85.4
28.5
0.3
0.6
0.0
1.8
3.4
6.7
0.9
43.2
72.8
1.3
0.3
0.4
False +/-
Eval @
1 ppm
+








+





+








+

Eval @
10 ppm
+




+


+
+







+





+


                                                                            i
'Sum of all  PAHs  detected.
                             4035 - 8
  Revision 0
January 1995

-------
 Table 6: Total PAH Content of Region X Field Samples Using EnSys
                     ,TM
             PAH RISc'M Immunoassay Test Kit (Contd.)

Sample ID
PAH-26
PAH-27
PAH- 2 8
PAH-29
PAH- 30
1 ppm Test
<1

*

*

>1




*
10 ppm Test
<10
*
*
*
*
*
>10






GC/MS
Lab Result
(ppm)1
27.9
0.0
16.4
0.4
9.6
False +/-
Eval @
1 ppm





Eval @
10 ppm
_

-


 Table 7: Total PAH Content of Region X Field Samples Using EnSys
  PAH RISc™ Immunoassay Test Kit Normalized to Cross-Reactivity

Sample ID
PAH-1
PAH-2
PAH -3
PAH -4
PAH -5
PAH- 6
PAH -7
PAH -8
PAH -9
PAH-10
PAH- 11
PAH-12
PAH-12Dup
PAH- 13
1 ppm Test
<1



*
*





*



>1
*




*


*
*




10 ppm Test
<10














>10
*
A
*


*
*
*
*
*

*
*
*

GC/MS
Lab Result
(ppm)1
0.1
8. 1
9.0
0.0
0.2
5.2
56.9
73.2
0.1
0.0
0.0
47.3
47.3
11.5
False +/-
Eval @
1 ppm
+







+
+




Eval @
10 ppm
+
+
+


+


+
+




xSum  of  all  PAHs  detected.
                             4035 - 9
  Revision 0
January 1995

-------
   Table 7: Total  PAH Content  of  Region X  Field  Samples  Using  EnSys
PAH RISc   Immunoassay Test Kit Normalized to Cross-Reactivity (Contd.)

Sample ID
PAH- 14
PAH- 15
PAH- 16
PAH- 17
PAH-18
PAH- 19
PAH- 20
PAH-21
PAH-22
PAH-23
PAH- 2 4
PAH-25
PAH-26
PAH- 2 7
PAH-28
PAH-29
PAH- 30
1 ppm Test
<1
it

it



*




*

*

*

>1

*

*
*
*



*
*





*
10 ppm Test
<10
it

it

*
*
*



*
*
*
it
*
*
*
>10



*



A
*
*








6C/MS
Lab Result
(ppm) '
0.2
0.5
0.0
1.2
1.7
3.6
0.6
27.5
49.2
0.8
0.1
0.2
13.5
0.0
6.4
0.2
2.8
False +/-
Eval @
1 ppm

+







+
+






Eval @
10 ppm



+





+


_




   'Sum of all PAHs detected.
                               4035 - 10
  Revision 0
January 1995

-------
                                  METHOD 4040

                  SOIL SCREENING FOR TOXAPHENE  BY  IMMUNOASSAY
1.0   SCOPE AND APPLICATION

      1.1   Method 4040 is a procedure for  screening soils to determine whether
toxaphene (CAS Registry 8001-35-2)  is  present at concentrations above 0.5 jug/g.
Method 4040 provides an estimate for the concentration of toxaphene by comparison
against standards.

      1.2   In cases  where  the exact concentration of  toxaphene  is required,
additional  techniques   (i.e.,   gas  chromatography  (Method  8081)   or  gas
chromatography/mass spectrometry (Method 8270)) should be used.

      1.3   This  method  is  restricted to  use  by  or under  the  supervision of
trained  analysts.   Each analyst  must  demonstrate  the  ability   to  generate
acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Test  kits  are   commercially  available  for  this  method.    The
manufacturer's directions should be followed.

      2.2   In general,  the method  is  performed  using  an  extract of  a soil
sample.  Filtered extracts may be stored cold,  in  the dark.   An aliquot of the
extract  and an enzyme-toxaphene  conjugate reagent  are added to  immobilized
toxaphene antibody.   The  enzyme-toxaphene  conjugate  "competes"  with toxaphene
present  in  the sample  for  binding  to  immobilized  toxaphene  antibody.   The
enzyme-toxaphene  conjugate  bound  to  the toxaphene  antibody then  catalyzes  a
colorless substrate to a colored product.  The test is interpreted by comparing
the color produced by  a sample to the response produced by a reference reaction.


3.0   INTERFERENCES

      3.1   Compounds that  are chemically similar may  cause a positive test
(false positive)  for  toxaphene.  The  test  kit  used  to  develop  this method was
evaluated for  interferences,  and  found to be relatively  insensitive  to other
organochlorine pesticides (e.g., Lindane, DDT and DDE).  The data.for the lower
limit of  detection  of these compounds are provided  in  Table 1.   Consult the
information  provided  by  the  manufacturer of  the  kit used  for  additional
information regarding cross  reactivity with other  compounds.

      3.2   Storage and  use temperatures  may  modify the method  performance.
Follow the manufacturer's directions for storage and use.
                                    4040-1                          Revision 0
                                                                  January 1995

-------
4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay  test  kit: EnviroGard™  Toxaphene in  Soil  (Millipore,
Inc.), or  equivalent.    Each  commercially available  test  kit will  supply or
specify the apparatus and materials necessary for successful completion of the
test.
5.0   REAGENTS

      5.1   Each  commercially  available test  kit  will  supply or  specify the
reagents necessary for successful completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory material  to  this  chapter,  Organic Analytes,
Section 4. 1 .

      6.2   Soils samples may be contaminated, and should therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions  for the test kit being used.
Those test kits used must meet or exceed the performance specifications indicated
in Tables 2-5.
8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit being used
for quality control  procedures  specific to the test  kit  used.   Additionally,
guidance provided in Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly  when  results  indicate
concentrations  near  the  action  level,  is  recommended  to refine  information
gathered with the kit.

      8.3   Do not use test kits past their expiration date.

      8.4   Do not use tubes or reagents  designated for use with  other test kits.

      8.5   Use the  test  kits  within their specified  storage  temperature and
operating temperature limits.

      8.6   Method 4040  is intended for field or laboratory use.  The appropriate
level  of quality  assurance should  accompany the  application  of this  method to
document data quality.


                                    4040-2                          Revision 0
                                                                  January 1995

-------
9.0   METHOD PERFORMANCE

      9.1   A  single  laboratory study  was conducted  using  five  spiked  soil
samples and 30 real-world soils contaminated with Toxaphene.  Results using the
immunoassay kit  were  compared  with  results  obtained  using  Method 3540/8081.
Table 2 presents the data from this study.  These data show a positive bias in
the  immunoassay  of almost  18%, compared to  a  14%  negative  bias  in  Method
3540/8081.

      9.2   Table 3 provides the results of single determinations in soil  from
New Mexico.

      9.3   Tables 8 and 9 provide data on the precision of Method 4040.


10.0  REFERENCES

1.   EnviroGard™  Toxaphene  Soil  Test  Kit  Guide, Millipore,  Inc.

2.   Marsden,  P.J.,  S-F  Tsang,  V.  Frank, N.  Chau, and M. Roby "Comparison of the
    Millipore  Immunoassay for Toxaphene with Soxhlet Extraction and Method 8081",
    Science Applications International Corporation, under contract to Millipore
    Inc.,  May  1994,  unpublished.
                                    4040-3                          Revision 0
                                                                  January 1995

-------
TABLE 1. POSSIBLE SOIL INTERFERENCES (a)
Compound
Diesel fuel
Endrin
Endosulfan I
Endosulfan II
Dieldrin
Heptachlor
Aldrin
technical Chlordane
gamma-BHC (Lindane)
alpha-BHC
delta-BHC
Soil Equivalent Concentrat
(ppb) Required to Yield a
Result
ion /xg/kg
Positive
45000
6
6
6
6
6
20
14
300
1000
1000
The following compounds were found to yield a negative result for concentrations
up to 200,000 jug/kg:
            Gasoline
            Pentachlorophenol
            DDT
PCB (Aroclor 1248)
Trinitrotoluene
(a)  Millipore, Inc. product literature
                                    Table  2
                Comparison of Method 4040 and Method 3540/8081

Spike
Concentration
(M9/9)
0.25
0.50
1.0
2.5
5.0
EnviroGard™ Toxaphene in Soil
Results
(/*g/g)
0.27
0.66
1.02
2.8
6.7
Average
% Recovery
108
132
102
112
134
117.6
Method 3540/Method 8081
Results
(/-tg/g)
0.19
0.33
0.83
2.9
5.5

% Recovery
75
66
84
116
110
L 85.6
                                    4040-4
                                Revision 0
                              January 1995

-------
             Table 3.
Results for New Mexico Soil  Samples
Sample ID |f Toxaphene (jug/g)
Lab 1
Ml
M2
M3
M4
M5*
M6*
M7
LMS12
M8*
M9
M10
Mil*
M12*
M13*
M14
M15*
M16
M16-MS
M16-MSD
M17
Lab 2 II Method 3540/8081
II (Lab 1)
28,89
70,104
54,89
103,50
10,30
45,33
Nazilini
soil #12
0,33
23,104
78,33
64,5
53,75
33,75
17,75
65,33
82,75
82,75
82,75
19,50
0.09 J
0.04 J
0.04 J
0.01 J
40
19.3
<0.50
<0.50
0.26 J
1.0
0.14 J
0.27 J
27.2
0.14 J
0.48 J
0.21 J
6.0
6.0
4.8
Method 4040
(Lab 2)
0.3
0.9
0.8
0.2
58
54.8
0.2
1.7
1.1
2.6
2.1
11
42
0.9
2.8
1.8
NA
NA
L_ 6

AGREEMENT3
Y, FN, FP
FP
FP
FP
FP
Y
FP
Y
FP
FP
FP
FP
FP
Y
FP
FP
FP
-
-
Y
              4040-5
  Revision 0
January 1995

-------
                             Table  3.  (continued)
Sample ID
Lab 1
M18
M19
M20
M21
M22
M23*
M24*
M24-MS
M24-MSD
M25*
M26
M27
M28
M29
M30*
M31
Lab 2
97,104
48,104
0,50
102,75
84,50
25,33
0,75
0,75
0,75
12,40 pit
0,89
0,104
98,89
104,33
76,89
40,50
Toxaphene (/j.g/g)
Method 3540/8081
(Lab 1)
0.049 J
0.054 J
1.3
0.15 J
0.058 J
89.6
0.5
3.7
3.6
35.6
0.16 J
0.88
0.41 J
0.30 J
0.10 J
323
Method 4040
(Lab 2)
0.6
1.1
2.3
0.3
0.9
130
1.9
NA
NA
45.5
6.9
2.1
3.4
0.7
5.8
460

AGREEMENT3
Y, FN, FP
FP
FP
Y
Y
FP
Y
FP
-
-
Y
FP
FP
FP
FP
FP
Y
NA = not analyzed

J = an  estimate  value.   This is used to indicate the result  is  less  than  the
sample quantitation limit but greater than zero.

* DDE identified using GC/MS analyses
                                    4040-6
  Revision 0
January 1995

-------
                                    Table 4
               Optical Measurement Precision for Spiked Samples
Spike Level
(ppm)
0.40
4.0
0.40
4.0
0.40
4.0
Mean O.D.
(450 nm)
0.798
0.450
0.753
0.397
0.713
0.385
Percent RSD
5.6
9.7
9.0
4.7
6.6
7.8
n
8
8
8
8
8
L_ 8
O.D.
(2 ppm spike)
0.540
0.540
0.501
0.501
0.541
0.541
                                    Table  5
             Recovery and Precision of Three Types  of Spiked Soils
Spike Cone, (ppm)
1.0
10.0
1.0
10.0
1.0
10.0
Mean percent
recovery
91.1
96.9
84.1
89.4
122.4
101.7
Percent RSD
20.0
10.4
14.6
4.2
8.8
2.0
n
3
3
3
3
3
3
overall percent recovery (n = 9), 1 ppm = 99 + 16%
overall percent recovery (n = 9), 10 ppm= 96 +  5%
                                    4040-7
  Revision 0
January 1995

-------
                                  METHOD 4041

                  SOIL SCREENING FOR CHLORDANE  BY  IMMUNOASSAY

1.0   SCOPE AND APPLICATION

      1.1   Method 4041 is a procedure for  screening soils to determine whether
chlordane (CAS Registry 57-74-9)  is present at concentrations above 20, 100 or
600 jug/kg.  Method 4041 provides an estimate for the concentration of chlordane
by comparison against standards.

      1.2   In cases  where  the exact concentration of  chlordane  is required,
additional  techniques   [i.e.,   gas  chromatography  (Method   8081)   or  gas
chromatography/mass spectrometry  (Method 8270)] should be used.

      1.3   This  method  is  restricted to  use  by  or under  the  supervision of
trained  analysts.   Each analyst must  demonstrate  the  ability to  generate
acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Test  kits  are   commercially  available  for  this  method.    The
manufacturer's directions should  be followed.

      2.2   In general,  the method  is  performed  using  an  extract  of  a  soil
sample.  Filtered extracts may be stored cold,  in the dark.   An aliquot of the
extract  and  an enzyme-chlordane  conjugate reagent  are added  to  immobilized
chlordane antibody.   The  enzyme-chlordane  conjugate  "competes"  with chlordane
present in the sample for binding to chlordane antibody.  The enzyme-chlordane
conjugate bound to the chlordane  antibody then catalyzes a colorless substrate
to a colored product.  The test is interpreted by comparing the color produced
by a sample to the response produced by a reference reaction.


3.0   INTERFERENCES

      3.1   Compounds that  are chemically similar  may  cause a  positive  test
(false positive)  for  chlordane.   The  test  kit  used  to develop this  method was
evaluated for interferences.  The  data for the  lower  limit of detection of these
compounds are  provided  in Table  1.   Consult  the information  provided  by the
manufacturer  of  the  kit  used  for  additional   information  regarding  cross
reactivity with other compounds.

      3.2   Storage and  use temperatures  may  modify the method  performance.
Follow the manufacturer's directions for storage and use.
                                    4041-1                          Revision 0
                                                                  January 1995

-------
4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay  test  kit: EnviroGard™  Chlordane in  Soil  (Millipore,
Inc.), or  equivalent.    Each  commercially available  test  kit will  supply or
specify the apparatus and materials necessary for successful completion of the
test.


5.0   REAGENTS

      5.1   Each  commercially  available test kit  will  supply or  specify the
reagents necessary for successful completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory material  to this chapter,  Organic  Analytes,
Section 4.1.

      6.2   Soil samples may be contaminated, and should  therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions  for the test kit being used.
Those test  kits  used must meet or  exceed the performance specifications indicated
in Tables 2-5.
8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit being used
for quality control  procedures  specific to the test  kit  used.   Additionally,
guidance provided in Method 4000 and Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly  when  results  indicate
concentrations  near  the  action  level,  is  recommended to refine  information
gathered with the kit.

      8.3   Do not use test kits past their expiration date.

      8.4   Do not use tubes or reagents  designated for use with  other test kits.

      8.5   Use the  test  kits  within their specified  storage  temperature and
operating temperature limits.

      8.6   Method 4041  is intended for field or laboratory use.  The appropriate
level of quality  assurance  should  accompany the  application  of this method to
document data quality.


                                    4041-2                          Revision 0
                                                                  January 1995

-------
9.0   METHOD PERFORMANCE

      9.1   Method sensitivity was determined by establishing  the  "noise"  level
expected from matrix effects encountered in negative soil samples and determining
the corresponding Chlordane concentration by comparison to the  analyte-specific
response curve. 8 different soils which did  not  contain Chlordane  were  assayed.
Each of these soils was extracted in triplicate  and each extract was  assayed  in
three different  assays.   The  mean and the standard deviation  of  the resulting
%Bo's (%Bo = [(ODsample/ODnegatlvecontrol)xlOO]) was calculated and  the sensitivity was
estimated at two standard deviations below the mean.  The sensitivity  for Method
4041 was determined to be 80% Bo at  a  95%  confidence  interval.   Based on the
average assay  response  to Chlordane,  this corresponds  to 6.4 ^g/kg  Chlordane.
These data are shown in Table 2.

      9.2   The  effect  of water  content of  the  soil  samples was  determined  by
assaying three different soil  samples which  had  been  dried and  subsequently had
water added to 30%  (w/w).   Aliquots of these samples were then fortified with
Chlordane  (100 ^g/kg).   Each soil  sample  was  assayed  three  times, with and
without added  water,   and  with  and  without  Chlordane   fortification.   It was
determined that water in soil  up to 30% had  no detectable effect on the method.
These data are shown in Table 3.

      9.3   The effect of the  pH of  the soil  extract was determined by adjusting
the soil pH of  three soil  samples.   Soil samples were adjusted to pH 2 - 4  using
6N HC1 and pH  10 - 12 using 6N NaOH.  Aliquots of the  pH adjusted soil samples
were  fortified  with  Chlordane  (100  ^g/kg).   Each soil  sample was  assayed
unadjusted and  with pH adjusted to 2-4 and  10-12,  both  unfortified and  fortified.
It  was  determined that  soil  samples with  pH  ranging  from   3  to   11  had   no
detectable effect on  the performance of the method.   These data  are shown  in
Table 4.

      9.4   A  field  trial was   undertaken  to  evaluate  to   ability  of the
EnviroGard™ Chlordane in  Soil Test  Kit to identify chlordane contaminated soil
at a remediation site.   A total of 32 soil samples were  evaluated by both Method
4041 and Method 8080.  Interpretation of the  results at a 1 mg/kg cutoff resulted
in 2/32  (6.3%)  false negatives and 0/32 (0%)  false positives.   Interpretation  of
the results at a cutoff of 10 mg/kg resulted in 0/32 (0%)  false  negatives and
2/32 (6.3%) false positives.  These data are shown in Table 5.

10.0  REFERENCES

1.    EnviroGard™ Chlordane in  Soil Test Kit Guide,  Millipore, Inc.
                                    4041-3                          Revision 0
                                                                  January 1995

-------
TABLE 1. CROSS REACTIVITY (a)
Compound
Chlordane
Endrin
Endosulfan I
Endosulfan II
Dieldrin
Heptachlor
Aldrin
Toxaphene
gamma-BHC (Lindane)
alpha-BHC
delta-BHC
Concentration Required for
Positive Interpretation (/*g/kg)
5
3
3
3
3
3
10
100
300
1000
1000
The following compounds were found to yield a negative result for
concentrations up to 200,000 jug/kg:
Gasoline PCB (Aroclor 1248)
Pentachlorophenol Trinitrotol uene
                                                i
4041-4
  Revision 0
January 1995

-------
AVERAGE
                                    TABLE  2
                              METHOD SENSITIVITY
Soil#
SI
S2
S3
S4
S5
S6
S7
S8
Part 1
Soil Type
LOAM/SAND
LOAM
CLAY
CLAY
CLAY
LOAM/SAND
SAND
SAND
- Average Response with
Average %Bo (n = 8)
92.8
86.2
85.5
95.4
83.9
88.5
81.4
95.8
Negative Soils
Standard Deviation
2.0
1.0
8.8
1.1
2.6
1.8
2.7
0.8
88.7
4.5
             Part 2 - Average Response with Chlordane Calibrators
     Chlordane
Concentration
  Average Absorbance    Average %Bo
0
5
25
125
500
1.043
0.882
0.598
0.322
0.159
N/A
84.4
57.2
30.8
15.2
                          Part  3  - Method  Sensitivity

Based on Part 1 and Part 2 Above:
         Average %Bo - 2 SD = 79.7 which is equivalent to 6.4 ^g/kg Chlordane
         Average %Bo - 3 SD = 75.2 which is equivalent to 8.6 ^g/kg Chlordane
                                    4041-5
                                      Revision 0
                                    January 1995

-------
                TABLE 3
EFFECT OF WATER CONTENT IN SOIL SAMPLES
Soil a
si
si
SI
SI
S2
S2
S2
S2
S3
S3
S3
S3
', Water
0
30
0
30
0
30
0
30
0
30
0
30
Fortified? Reo. 1
No
No
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
95.2
96.0
40.5
42.2
85.8
78.7
37.7
39.8
76.6
87.4
40.0
40.8
Reo. 2
101
99.2
38.5
43.0
87.1
84.9
39.5
38.8
76.6
88.7
40.2
37.1
Rep. 3
94.5
96.0
35.9
43.0
85.5
79.8
40.6
37.0
73.0
85.7
35.7
38.7
Mean Std. Dev.
97.0
97.1
38.3
42.8
86.1
81.1
39.3
38.5
75.4
87.3
38.7
38.9
3.7
1.8
2.3
0.5
0.9
3.3
1.5
1.4
2.1
1.5
2.5
1.9
± 2 SD Ranqe
89.6
93.5
33.7
41.8
84.3
74.5
36.3
35.7
71.2
84.3
33.7
35.1
- 104
- 101
- 42.9
- 43.8
- 87.9
- 87.8
- 42.3
- 41.3
- 79.6
- 90.3
-43.7
- 42.7
                 4041-6
  Revision 0
January 1995

-------
                                    TABLE  4
                         EFFECT OF  pH  OF SOIL SAMPLES
oil
SI
SI
SI
SI
SI
SI
S2
S2
S2
S2
S2
S2
S3
S3
S3
S3
S3
S3
pH Adj. Fortified? Rep
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
97
97
107
43
43
44
87
94
89
43
44
42
72
85
89
44
40
40
. 1* Rep. 2
.5
.4

.6
.6
.8
.8
.2
.3
.9
.6
.3
.3
.3
.4
.5
.5
.6
87.8
114
114
47.5
51.8
50.8
86.3
108
99.3
48.9
55.9
59.2
74.5
105
83.8
49.5
52.1
37.1
Rep. 3
94.8
94.7
93.8
38.5
34.1
32.0
87.6
80.5
77.9
33.9
41.5
36.5
79.3
75.7
85.9
32.6
34.7
43.9
Mean
93.4
102
105
43.2
43.2
42.5
87.3
94.1
88.8
42.2
47.4
46.0
75.4
88.8
86.4
42.2
42.4
40.5
Std.
5
10
10
4
8
9
0
13
10
7
7
11
3
15
2
8
8
3
Dev.
.0
.7
.1
.5
.8
.6
.8
.5
.7
.7
.6
.8
.6
.1
.8
.7
.9
.4
± 2 SD Range
83.4
80.8
84.7
34.2
25.6
23.3
85.7
67.1
67.4
26.8
32.2
22.4
68.2
58.6
80.8
24.8
24.6
33.7
- 103
- 124
- 125
- 52.2
- 60.8
- 61.7
- 88.9
~ 121
- 110
- 57.6
- 62.6
- 69.6
- 82.6
- 119
- 92.0
- 59.6
- 60.2
- 47.3
All values  shown are %Bo [= (ODsample/ODnegatlve C0ntrol)xl00]
                                    4041-7
  Revision  0
January  1995

-------
Sample ID

co-ss-2
co-ss-3
co-ss-4
co-ss-5
co-ss-6

co-ss-7
co-ss-8
co-ss-9
co-ss-10
co-ss-13

co-ss-14
co-ss-15
co-ss-17
co-ss-20
co-ss-21

co-ss-22
co-ss-23
co-ss-24
co-ss-25
co-ss-26

co-ss-27
co-ss-28
co-ss-28-170
co-ss-29
co-ss-30

co-ss-31
co-ss-32
co-ss-33
co-ss-34
co-ss-35

co-ss-36
co-ss-41
                      TABLE 5
            Correlation to Method 8081

          Test  Interpretation at 1 mg/kg

Method 8081 (mg/kg)      Immunoassay  (mg/kg)
        45
         4.9
        25
         1.4
         2.7

         2.5
        <1.0
         7.9
         6.0
         5.2

         2.9
         2.1
        <1.0
         2.8
        51
         1.4
         9.6
         2.6
        14
         1.8
         2.9
         4.2
         5.9
POSITIVE
POSITIVE
POSITIVE
NEGATIVE
POSITIVE

POSITIVE
NEGATIVE
POSITIVE
POSITIVE
POSITIVE

POSITIVE
POSITIVE
NEGATIVE
NEGATIVE
NEGATIVE

NEGATIVE
NEGATIVE
POSITIVE
POSITIVE
NEGATIVE

NEGATIVE
POSITIVE
POSITIVE
POSITIVE
POSITIVE

NEGATIVE
NEGATIVE
POSITIVE
POSITIVE
NEGATIVE

POSITIVE
NEGATIVE
Results Agree?

      YES
      YES
      YES
FALSE NEGATIVE
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
FALSE NEGATIVE
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
                                    4041-8
                                                      Revision 0
                                                    January 1995

-------
                              TABLE 5 (continued)
                          Correlation to Method 8081

                        Test Interpretation  at  10  mg/kg
Sample ID     Method 8081 (mg/kg)
                 Immunoassay  (mg/kg)     Results Agree?
co-ss-2
co-ss-3
co-ss-4
co-ss-5
co-ss-6

co-ss-7
co-ss-8
co-ss-9
co-ss-10
co-ss-13

co-ss-14
co-ss-15
co-ss-17
co-ss-20
co-ss-21

co-ss-22
co-ss-23
co-ss-24
co-ss-25
co-ss-26

co-ss-27
co-ss-28
co-ss-28-170
co-ss-29
co-ss-30

co-ss-31
co-ss-32
co-ss-33
co-ss-34
co-ss-35

co-ss-36
co-ss-41
45
 4.9
25
 1.4
 2.7

 2.5
<1.0
 7.9
 6.0
 5.2

 2.9
 2.1
<1.0
 2.8
51
 1.4
 9.6
 2.6
14
 1.8
 2.9
 4.2
 5.9
POSITIVE
NEGATIVE
POSITIVE
NEGATIVE
NEGATIVE

NEGATIVE
NEGATIVE
POSITIVE
POSITIVE
NEGATIVE

NEGATIVE
NEGATIVE
NEGATIVE
NEGATIVE
NEGATIVE

NEGATIVE
NEGATIVE
POSITIVE
NEGATIVE
NEGATIVE

NEGATIVE
NEGATIVE
NEGATIVE
POSITIVE
NEGATIVE

NEGATIVE
NEGATIVE
NEGATIVE
NEGATIVE
NEGATIVE

NEGATIVE
NEGATIVE
      YES
      YES
      YES
      YES
      YES

      YES
      YES
FALSE POSITIVE
FALSE POSITIVE
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
                                    4041-9
                                              Revision  0
                                            January  1995

-------
                                  METHOD 4042

                     SOIL SCREENING FOR DDT BY IMMUNOASSAY
1.0   SCOPE AND APPLICATION

      1.1   Method 4042 is a procedure for screening  soils  to determine whether
l,l,l-trichloro-2,2-bis (4-chlorophenyl)  ethane  (DDT) (CAS Registry 50-29-3) and
its breakdown products (ODD, DDE, and  DDA) are present at  concentrations above
0.2,  1.0 or  10 mg/kg.    Method 4042  provides an  estimate  for the  sum  of
concentrations of DDT and daughter compounds by comparison against standards.

      1.2   In cases where the exact concentration of DDT is required, additional
techniques  [i.e., gas  chromatography  (Method 8081)  or gas chromatography/mass
spectrometry (Method 8270)] should be  used.

      1.3   This method  is  restricted to  use  by  or under the supervision  of
trained  analysts.    Each  analyst  must  demonstrate  the  ability  to  generate
acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Test  kits  are  commercially  available  for   this method.    The
manufacturer's directions should be followed.

      2.2   In general,  the method  is performed  using  an extract of  a soil
sample.   Filtered extracts may be stored cold, in  the dark.  An aliquot of the
extract  and an  enzyme-DDT conjugate  reagent  are  added  to  immobilized  DDT
antibody. The enzyme-DDT conjugate "competes" with DDT present in the sample for
binding to DDT antibody.  The enzyme-DDT conjugate bound to the DDT  antibody then
catalyzes a colorless substrate to a colored product.  The test is interpreted
by  comparing  the color  produced by a sample to  the  response produced  by  a
reference reaction.
3.0   INTERFERENCES

      3.1   Compounds that  are chemically similar  may cause a  positive test
(false positive)  for DDT.  The test kit used to develop this  method was evaluated
for interferences.   The data for the lower limit  of detection of these compounds
are provided in Table 1.   Consult the information provided by the manufacturer
of the kit  used for additional information regarding  cross reactivity with other
compounds.

      3.2   Storage  and  use temperatures may  modify the  method  performance.
Follow the manufacturer's directions for storage and use.
                                    4042-1                          Revision 0
                                                                  January 1995

-------
4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay test kit:  EnviroGard™ DDT  in Soil (Millipore, Inc.), or
equivalent.  Each  commercially  available test kit will  supply  or  specify the
apparatus and materials necessary for successful  completion of the test.


5.0   REAGENTS

      5.1   Each commercially  available  test kit  will  supply  or  specify the
reagents necessary for successful  completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Section 4.1.

      6.2   Soil samples may be contaminated, and  should therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions for  the test  kit being used.
Those test kits  used must meet or exceed the performance  specifications indicated
in Tables 2-5.
8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit being used
for quality control  procedures  specific to the test  kit  used.   Additionally,
guidance provided in Method 4000 and Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly  when  results  indicate
concentrations  near  the  action  level,   is  recommended to refine  information
gathered with the kit.

      8.3   Do not use test kits past their expiration date.

      8.4   Do not use tubes or reagents  designated for use with  other test kits.

      8.5   Use the  test  kits  within their specified  storage  temperature and
operating temperature limits.

      8.6   Method 4042  is intended for field or laboratory use.  The appropriate
level  of quality  assurance  should  accompany the  application  of this method to
document data quality.
                                    4042-2                          Revision 0
                                                                  January 1995

-------
9.0   METHOD PERFORMANCE

      9.1   Method sensitivity was determined by establishing  the  "noise"  level
expected from matrix effects encountered in negative soil samples and determining
the  corresponding DDT  concentration  by comparison  to  the   analyte-specific
response curve.  8 different  soils which  did not contain DDT were assayed.   Each
of these soils  was extracted  in triplicate and each extract was assayed  in  three
different assays.  The  mean  and  the standard deviation of the  resulting  %Bo's
(%Bo  =  [(ODsample/ODnegativecontrol)xlOO])  were calculated  and  the  sensitivity was
estimated at two standard deviations  below the mean.  The sensitivity  for Method
4042 was determined to  be  81.4%  Bo  at a 95% confidence inteval.  Based on the
average assay response to DDT, this corresponds to 0.044  ppm DDT.  These  data are
shown in Table 2.

      9.2   The  effect  of  water  content  of  the soil  samples was determined  by
assaying three different soil samples which  had  been dried and  subsequently had
water added to  30% (w/w).  Aliquots of these  samples were then fortified  with DDT
(1.0 mg/kg).  Each soil  sample was  assayed  three  times, with and without  added
water, and with and without DDT fortification.  It was determined  that  water  in
soil up to 30%  had no  detectable  effect on the method.   These data are  shown  in
Table 3.

      9.3   The effect of the pH  of  the soil  extract was  determined by adjusting
the soil pH of  three soil  samples.   Soil  samples were  adjusted to pH 2 - 4  using
6N HC1 and pH 10  - 12 using 6N NaOH.  Aliquots of the pH  adjusted soil samples
were fortified  with DDT  (1.0 mg/kg).  Each soil sample was assayed unadjusted and
with  pH  adjusted to  2-4  and 10-12,  both unfortified  and fortified.   It was
determined that  soil  samples with  pH  ranging from  3 to  11  had no detectable
effect on the performance  of the method.  These data are  shown  in Table 4.

      9.4   A  field  study  was  conducted  at  a  contaminated  site   using   a
commercially available  test  kit  (EnviroGard™ DDT in Soil Test Kit,  Millipore
Corp.).   The immunoassay was used to  identify soil which  had been contaminated
with  DDT.   The  standard  method  (Method 8080)  was  performed  at  a  certified
laboratory and  the results were compared  to  the  immunoassay.  When interpreting
the  results  at  a 0.2  ppm cutoff,  the   immunoassay yielded   0/32  (0%)   false
negatives and 2/32 (6.3%)  false positives.  When  interpreting  the results  at  a
1.0 ppm cutoff,  the  immunoassay  yielded  1/32  (3.1%) false negatives  and  2/32
(6.3%) false positives.   These data are  shown in Table  5.


10.0  REFERENCES

1.  EnviroGard™ DDT  in  Soil  Test Kit Guide, Millipore,  Inc.
                                    4042-3                          Revision 0
                                                                  January 1995

-------
TABLE 1. CROSS REACTIVITY (a)
Compound
p,p'-DDT
p,p'-DDD
p,p'-DDE
o,p'-DDT
o,p'-DDD
o,p'-DDE
DDA
Chloropropylate
Chlorobenzilate
Dicofol
Chloroxuron
Monol inuron
Thiobencarb
Tebuconazole
Neburon
Tetradifon
Diclofop
PCB (Aroclor 1248)
Concentration Required for
Positive Interpretation (ppm)
0.04
0.01
0.18
4.0
0.4
3.0
0.002
0.007
0.03
0.14
24
25
5
7
17
1.2
70
90
The following analytes are not detected at or above 100 ppm:
2,4-D 4-chlorophenoxyacetic acid Chlordane
Pentachlorophenol Chlorbromuron Chlortoluron
Dicamba Diflubenzuron Diuron
Lindane Linuron MCPA acid
MCPB Mecoprop Gasoline
Diesel Fuel 2,4,6-Trinitrotoluene Toxaphene
                                                i
4042-4
  Revision 0
January 1995

-------
AVERAGE
                                    TABLE 2
                              Method Sensitivity
Soil#
SI
S2
S3
S4
S5
S6
S7
S8
Part 1
Soil Type
LOAM
CLAY
SAND
LOAM
LOAM/SAND
CLAY
LOAM/SAND
SAND/ LOAM
- Average Response with
Average %Bo (n = 9}
87.0
93.2
97.2
87.7
88.1
100.8
103.6
89.6
Negative Soils
Standard Deviation
7.5
2.3
2.6
1.2
2.3
2.1
0.3
4.5
93.4
6.0
                Part 2 - Average Response with DDT Calibrators
      DDT
Concentration (ppm)
Average Absorbance
Average %Bo
0
0.1
1.0
10.0
50.0
1.133
0.897
0.569
0.362
0.259
N/A
79.4
50.3
32.0
22.9
                          Part 3  -  Method  Sensitivity

Based on Part 1 and Part 2 Above:
Average %Bo - 2 SD = 81.4 which is equivalent to 0.044 ppm DDT
Average %Bo - 3 SD = 75.4 which is equivalent to 0.097 ppm DDT
                                    4042-5
                                       Revision 0
                                     January 1995

-------
                                   TABLE 3
                   EFFECT OF WATER CONTENT  IN SOIL SAMPLES
SI
SI
SI
SI

S2
S2
S2
S2

S3
S3
S3
S3
 0
30
 0
30

 0
30
 0
30

 0
30
 0
30
•tified?
No
No
Yes
Yes
No
No
Yes
Yes
No
No
Yes
Yes
Rep. 1
79.7*
89.1
49.8
55.8
85.2
94.8
54.4
56.3
96.2
95.6
54.8
59.4
Rep
79
84
62
59
96
94
47
53
91
90
52
55
. 2
.3
.0
.1
.9
.2
.3
.0
.8
.3
.5
.9
.0
Rep
83
85
46
58
97
95
56
60
100
96
54
54
. 3
.7
.9
.3
.0
.9
.0
.1
.2
.0
.4
.8
.5
Mean
80.9
86.4
52.8
57.9
93.1
94.7
52.5
56.8
95.8
94.2
54.2
56.3
Std.
2
2
8
2
6
0
4
3
4
3
1
2
Dev.
.4
.6
.3
.1
.9
.3
.8
.2
.3
.2
.1
.7
± 2
76.
81.
36.
53.
79.
94.
42.
50.
87.
87.
52.
50.
SD
1 -
2 -
2 -
7 -
3 -
1 -
9 -
4 -
2 -
8 -
0 -
9 -
Range
85.7
91.6
69.4
62.1
106.9
95.3
62.1
63.2
104.4
100.6
56.4
61.7
* All values  shown are %Bo [=  (ODsample/ODn
                                          egative conti
                                    4042-6
                                                               Revision 0
                                                             January 1995

-------
                                      TABLE 4
                           EFFECT OF pH OF SOIL  SAMPLES
oil
SI
SI
SI
SI
SI
SI
S2
S2
S2
S2
S2
S2
S3
S3
S3
S3
S3
S3
pH Adj. Fortified? Rep
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
None
Acidic
Basic
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
No
No
No
Yes
Yes
Yes
91
79
80
57
54
52
94
87
87
51
52
52
99
86
94
56
54
54
. 1* Rep. 2
.4
.7
.5
.5
.2
.8
.7
.8
.9
.7
.2
.0
.1
.4
.9
.2
.5
.6
91.3
87.0
84.5
60.3
60.6
60.2
90.6
100.1
81.6
56.9
61.0
53.5
94.2
84.3
100.3
54.3
53.5
57.2
Rep. 3
78.3
86.8
78.5
55.1
55.2
53.3
94.5
100.9
98.3
48.3
55.2
48.9
98.2
85.5
92.9
52.8
53.9
62.9
Mean Std.
87.0
84.5
81.2
57.6
56.7
55.5
93.2
96.3
89.3
52.3
56.1
51.5
97.2
85.4
96.1
54.4
54.0
58.2
7
4
3
2
3
4
2
7
8
4
4
2
2
1
3
1
0
4
Dev
.5
.1
.0
.6
.4
.1
.3
.3
.5
.3
.5
.3
.6
.1
.8
.7
.5
.2
. ± 2 SD Range
72.0
76.3
75.2
52.4
49.9
47.3
88.6
81.7
72.3
43.7
47.1
46.9
92.0
83.2
88.5
51.0
53.0
49.8
- 102
- 92.7
- 87.2
- 62.8
-63.5
- 63.7
- 97.8
- Ill
- 106
- 60.9
- 65.1
- 56.1
- 102
- 87.6
- 104
- 57.8
- 55.0
- 66.6
* All values  shown are %Bo  [=  (OD5amp,yODnegatlvecontrol)xlOO]
                                      4042-7                           Revision  0
                                                                     January  1995

-------
                                   TABLE  5
                           Comparison to Method 8081

                       Test Interpretation at 0.2 mg/kg
Sample ID

co-ss-2
co-ss-3
co-ss-4
co-ss-5
co-ss-6

co-ss-7
co-ss-8
co-ss-9
co-ss-10
co-ss-13

co-ss-14
co-ss-15
co-ss-17
co-ss-20
co-ss-21

co-ss-22
co-ss-23
co-ss-24
co-ss-25
co-ss-26

co-ss-27
co-ss-28
co-ss-28-170
co-ss-29
co-ss-30

co-ss-31
co-ss-32
co-ss-33
co-ss-34
co-ss-35

co-ss-36
co-ss-41
Method 8081 (mg/kg)

         3.6
         0.55
         2.3
        <0.05
         0.15

         0.3
         0.1
         0.8
         0.23
         0.79

         0.58
         0.35
        <0.05
         0.18
         0.06

        <0.05
        <0.05
         1.2
         0.12
        <0.05

        <0.05
         0.16
         0.18
         0.69
         0.73

         0.68
        <0.05
         0.32
         0.23
         0.52

         1.0
        <0.05
Immunoassay (mq/kg)

     POSITIVE
     POSITIVE
     POSITIVE
     NEGATIVE
     POSITIVE

     POSITIVE
     NEGATIVE
     POSITIVE
     POSITIVE
     POSITIVE

     POSITIVE
     POSITIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
     POSITIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
     POSITIVE
     POSITIVE
     POSITIVE

     POSITIVE
     NEGATIVE
     POSITIVE
     POSITIVE
     POSITIVE

     POSITIVE
     NEGATIVE
Results Agree?

      YES
      YES
      YES
      YES
FALSE POSITIVE

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
FALSE POSITIVE
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
                                    4042-8
                                                       Revision  0
                                                     January 1995

-------
                              TABLE  5  (continued)
                           Comparison  to  Method  8081

                       Test Interpretation at 1.0 mg/kg
Sample ID

co-ss-2
co-ss-3
co-ss-4
co-ss-5
co-ss-6

co-ss-7
co-ss-8
co-ss-9
co-ss-10
co-ss-13

co-ss-14
co-ss-15
co-ss-17
co-ss-20
co-ss-21

co-ss-22
co-ss-23
co-ss-24
co-ss-25
co-ss-26

co-ss-27
co-ss-28
co-ss-28-170
co-ss-29
co-ss-30

co-ss-31
co-ss-32
co-ss-33
co-ss-34
co-ss-35

co-ss-36
co-ss-41
Method 8081 (mg/kg)

         3.6
         0.55
         2.3
        <0.05
         0.15

         0.3
         0.1
         0.8
         0.23
         0.79

         0.58
         0.35
        <0.05
         0.18
         0.06

        <0.05
        <0.05
         1.2
         0.12
        <0.05

        <0.05
         0.16
         0.18
         0.69
         0.73

         0.68
        <0.05
         0.32
         0.23
         0.52

         1.0
        <0.05
Immunoassay (mg/kg)

     POSITIVE
     NEGATIVE
     POSITIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
     POSITIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE
     POSITIVE

     POSITIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE
     NEGATIVE

     NEGATIVE
     NEGATIVE
Results Agree?

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
      YES

      YES
      YES
      YES
      YES
FALSE POSITIVE

FALSE POSITIVE
      YES
      YES
      YES
      YES

FALSE NEGATIVE
      YES
                                    4042-9
                                                      Revision 0
                                                    January 1995

-------
                                  METHOD 4050

               TNT EXPLOSIVES IN WATER AND SOILS  BY  IMMIJNOASSAY
1.0   SCOPE AND APPLICATION

      1.1   Method  4050  is  a  procedure  for  screening waters  and  soils  to
determine  when  trinitrotoluene   (TNT,   CAS   No.   118-96-7)   is  present  at
concentrations  above  0.5  mg/kg  in soil  and  5  jiig/L  in  water.    Method 4050
provides an estimate for the concentration  of TNT by comparison  with  a reference.

      1.2   Using the test kit from which this method was developed,  93% of soil
samples containing 0.25 ppm or less of TNT will produce a negative result, and
99+% of soil samples containing 1.0 ppm or  greater of TNT will produce a positive
result.  In addition, 93% of water samples  containing 2.5 ppb or less of TNT will
produce a negative result, and 99%+ of water samples containing 10 ppb or more
of TNT will produce a positive result.

      1.3   In  cases  where  the  exact concentrations  of  TNT are  required,
quantitative techniques (i.e., Method 8330) should be used.

      1.4   This method  is restricted to  use  by or under  the supervision  of
trained  analysts.    Each  analyst  must demonstrate  the  ability  to  generate
acceptable results with this method.
2.0   SUMMARY OF METHOD

      2.1   Test  kits  are  commercially  available
manufacturer's directions should be followed.
for  this  method.
The
      2.2   In general, the method is  performed  using  a diluted water sample or
an extract of a  soil  sample.   Samples  and  an enzyme-TNT  conjugate reagent are
added to immobilized  TNT antibody.  The enzyme-TNT conjugate  "competes" with TNT
present in the sample for binding to immobilized TNT antibody.  The enzyme-TNT
conjugate bound  to the TNT  antibody then  catalyzes  a  colorless substrate to a
colored product.  The test is interpreted by comparing the color produced by a
sample to the response produced by a reference reaction.
3.0   INTERFERENCES

      3.1   Chemically similar compounds  and compounds that might be expected to
be found  in  conjunction with  TNT  contamination  were tested  to determine the
concentration required to produce a positive test result.

            3.1.1  Table 1 provides the concentrations of compounds tested  with
      the D TECH test  kit that  are required to elicit  a positive  response at the
      MDL,  as  well  as  the  concentration  required  to  yield  50%  inhibition
                                    4050-1
               Revision 0
            January  1995

-------
      compared to the standard curve.


4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay test kit:  D TECH™ TNT (Strategic Diagnostics Inc.), or
equivalent.   Each  commercially  available test kit will supply  or  specify the
apparatus and materials necessary for successful  completion of the test.


5.0   REAGENTS

      5.1   Each commercially  available test kit  will  supply  or  specify the
reagents necessary,, for successful  completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION, AND TRANSPORTATION

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Section 4.1.

      6.2   Soil samples may be contaminated, and  should therefore be considered
hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions  for  the test  kit being used.
Those test kits used must meet or exceed the performance  specifications indicated
in Tables 3-6.


8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit being used
for quality control  procedures  specific to the test kit  used.   Additionally,
guidance provided in Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly when   results  indicate
concentrations  near  the action  level,   is  recommended to refine  information
gathered with the kit.

      8.3   Do not use test  kits past their expiration date.

      8.4   Do not  use tubes  or reagents designated  for use with other test kits.
Do not mix reagents from one kit lot with a different kit lot.

      8.5   Use the  test kits within their  specified  storage temperature and
operating temperature limits.
                                    4050-2                          Revision 0
                                                                  January 1995

-------
      8.6   Method 4050 is  intended for field or laboratory use. The appropriate
level of quality  assurance should  accompany the  application of this method to
document data quality.


9.0   METHOD PERFORMANCE

      9.1   Table  1  provides data  on  the minimum  concentrations  of possible
interferants and co-contaminants required to elicit a positive response in the
test kits evaluated.

      9.2   Twenty five soil  samples, known to not be contaminated  with TNT, were
extracted and analyzed using the DTECH TNT kit to determine the extent of soil
matrix effects on the performance of the  test kit.   The results are provided in
Table 2A,  and  show that false  positive  results  are not  attributable  to  soil
components.  Table 2B presents similar data generated from the analysis of thirty
uncontaminated water samples.

      9.3   Thirty  water  samples  and  thirty  soil  samples,  known   to  not  be
contaminated with TNT, were each spiked with TNT at one-half and two times the
MDL (0.25 and 1.0  ppm respectively).  These samples were analyzed  with the DTECH
TNT test kit to determine the error rate of the assay.  The  results are presented
in Tables 3A and 3B.

      9.4   Ten different soil  types, all  known to not be contaminated with TNT,
were spiked with  an acetone solution containing approximately 1.0  ppm TNT.  This
spiking solution  was later  quantitated  by  Method 8330 and  found to contain  0.77
ppm TNT.  The spiked soil  samples were analyzed three (3)  times with the DTECH
kit to determine the extraction efficiency of the method.  The data are presented
in Table 4.

      9.5   Table  5 presents the  results  of analysis  of  three soils spiked at
approximately 1 and 3 ppm TNT.   Each sample  was analyzed once using Method 8330
and ten times using the DTECH kit.

      9.6   Tables 6A and  6B present the  results of two field trials.  In each
trial, soil  samples were obtained at a West Coast  site from borings,  using a
split spoon technique.  The samples were homogenized by placing approximately six
cubic inches of soil into  a stainless steel  vessel and mixing for five minutes
with a stainless  steel trowel.   The soil  was aliquotted into two (2) six ounce
glass  bottles,  tested  on-site  using the   DTECH  method  and  transported  to
commercial  laboratories (one laboratory per  field  trial) for  analysis by Method
8330.  Table 6C presents the results of a  third party field trial, conducted by
the California Department  of Environmental Health Services.


10.0  REFERENCES

1.    D TECH™ TNT Users Guide  , SDI/EM Sciences  1994
                                    4050-3                          Revision 0
                                                                  January 1995

-------
2.    Mutter,L., G. Teaney, and J.W.Stave, "A Novel Field Screening System for
      TNT Using EIA", in Field Screening Methods for Hazardous Wastes and
      Toxic Chemicals, Vol 1, Proceedings of the 1993 U.S. EPA/A&WMA
      International Symposium, p.472, 1993.

3.    Teaney, G., J.Melby, L.Mutter and J.Stave, "A Novel Field Analytical
      Method for TNT", Proceedings of the American Association of Analytical
      Chemists, 1993.

4.    Haas, R.J., and B.P. Simmons, "Measurement of Trinitrotoluene (TNT) and
      Hexahydro-l,3,5-trinitro-l,3,5-triazine (RDX) in Soil by Enzyme
      Immunoassay and High Performance Liquid Chromatography (EPA Method
      8330)", California Environmental Protection Agency, Department of Toxic
      Substances Control, Hazardous Materials Laboratory, March, 1995.
                                    4050-4                          Revision 0
                                                                  January  1995

-------
                              TABLE 1
                          CROSS REACTANTS
                       D TECH™ TNT test  kit
COMPOUND
TNT (2,4,6-trinitrotoluene)
Tetryld
1, 3, 5-tri nitrobenzene
2-amino-4,6-dinitrotoluene
4-amino-2,6-dinitrotoluene
2,4-dinitrotoluene
2,6-diaminonitrotoluene
2-nitrophenol
4-nitrophenol
2,4-dinitrophenol
RDXd
HMXd
MDLa
(ppm)
0.5
3
4
13
>500
90
>500
>500
>500
>500
>500
>500
1C 50b
(ppm)
17
48
75
150
>500
390
>500
>500
>500
>500
>500
>500
% CROSS
REACTIVITY0
100
35
23
11
<1
4
<1
<1
<1
<1
<1
<1
The following compounds were not detected at or above 100 ppm:
Benzene Xylenes PCB 1254 Triazine
Ethyl benzene Toluene PCP
PAHs - an equal concentration mixture of:
Acenaphthene Acenaphthalene Anthracene
1,2-Benzanthracene Benzo(a)pyrene Benzo(b)fluoranthene
Benzo(ghi )perylene Benzo(k)fl uoranthene Chrysene
Dibenz(ah)anthracene Fluoranthene Fluorene
Indeno(123-cd)pyrene Naphthalene Phenanthrene
Pyrene
The Method Detection Limit (MDL) is defined as the lowest concentration
of compound that yields a positive test result.
The IC50 is  defined as  the concentration  of compound  required  to produce
a test response equivalent to 50% of the maximum response.
% Cross reactivity is determined by dividing the equivalent TNT
concentration by the actual compound concentration at IC50
Tetryl = methyl-2,4,6-trinitrophenylnitramine
RDX = hexahydro-l,3,5-trinitro-l,3,5-triazine
HMX = octahydro-l,3,5,7-tetranitro-l,3,5,7-tetrazocine
                              4050-5
  Revision 0
January 1995

-------
     TABLE 2A
SOIL MATRIX EFFECTS
Soil
133
101
100
102
106
107
109
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
127
128
130
Soil type
Avonburg Fine Sine Silt
Matapeake Silt Loam
Clay Loam
Sassafras Sand Loam
Evesboro Low Organic Sand
Pokomoke High OM Sand
Davidson Clay Loam
Shontic Casa Grande Sand
Casa Grande Clay Loam
Trix Sand Clay Loam
Trix Casa Grande Clay
Yolo Loam
Capay Silt Clay
Sycamore Silt Loam
Dennis Silt Loam
Grundy Silt Clay Loam
Luray Silt Clay Loam
Wooster Silt Loam
Vienna Loam
Opal Clay
Raub Silt Loam
Rockfield Silt Loam
Ci sne
Muscatine Loam
Sandy Brae

N/A
DE
DE
DE
DE
DE
GA
AZ
AZ
AZ
AZ
CA
CA
CA
KS
KS
OH
OH
SO
SD
IN
IN
IL
IL
DE
D TECH
RANGE (ppm)
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
       4050-6
  Revision 0
January 1995

-------
      TABLE 2B
WATER MATRIX EFFECTS
Water
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
Water type
Ground Water, Burlington
Well Water, Burlington
Surface Water #1, Houston
Unknown Creek, Dartmouth
City Well Water, Ontario
Pacific Ocean, Victoria
Surface Water, Harmony Woods
Adamsville River, Adamsville
Surface Water #2, Houston
Buttermilk Falls, White Haven
Main St Pond, Germantown
Hudson River, Germantown
Atlantic Ocean
Ground Water #1, Dover
Ground Water #2, Dover
Ground Water #3, Dover
Drinking Well Water,
Ground Water, Elsmere
Ground Water, Elsmere
Ground Water, Elsmere
Lab Sample 20643
Lab Sample 20645
Lab Sample 20659
Lab Sample 20826
Lab Sample 20827
Lab Sample 20843
Lab Sample 20850
Lab Sample 20848
Ground Water, Adrian
Ground Water, Adrian

IA
IA
TX
MA
CA
CA
DE
RI
TX
PA
NY
NY
NJ
DE
DE
DE
PA
DE
DE
DE
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
GA
GA
D TECH RANGE (ppm)
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
<5.0
       4050-7
  Revision 0
January 1995

-------
                                   TABLE 3A
            False Negative and False Positive Rates, Soil Matrix5
Spike Concentration
0.25 ppm
1 .0 ppm
False Positive Rate
7%
-
False Negative Rate
_
0%
a Thirty negative soils were spiked with TNT at one-half and two times the MDL
(0.25 and 1.0 ppm respectively).  These samples were analyzed with the DTECH TNT
test kit to determine the error rate of the assay.
                                   TABLE 3B
            False Negative and False  Positive Rates,  Water Matrix6
                 i
Spike Concentration
0.25 ppm
1.0 ppm
False Positive Rate
7%
100%
False Negative Rate
93%
0%
a Thirty negative water samples were spiked with TNT at one-half and two times
the MDL (0.25 and 1.0 ppm respectively).  These samples were analyzed with the
DTECH TNT test kit to determine the error rate of the assay.
                                    4050-8
  Revision 0
January 1995

-------
                                    TABLE  4
           DETERMINATION OF EXTRACTION EFFICIENCY FROM SOIL SAMPLES3
SOIL ID
101
106
108
109
110
116
117
123
126
128

SPIKING
SOLUTION
MEAN TNT CONC.
(ppm)
0.54
0.64
0.87
0.63
0.88
1.02
0.82
0.87
0.95
0.65

0.77
SD
0.04
0.06
0.18
0.08
0.15
0.15
0.15
0.23
0.26
0.11

N/A
%CV
7
9
20
13
17
17
15
26
28
16

N/A
%RECOVERY
70
84
113
82
115
115
132
113
123
84

100
aTen different TNT negative soils were spiked with an acetone solution containing
0.77 TNT.  The  spiked soil  samples  were  analyzed  three times with the DTECH kit
to determine the extraction efficiency of the method.
                                    4050-9
  Revision 0
January 1995

-------
                                    TABLE  5
                    RECOVERY OF TNT SPIKED INTO REAL SOILS
Three (3) soils were spiked at approximately 1 and 3 ppm TNT.
analyzed once by Method 8330 and ten (10) times by D TECH.
Each sample was
SAMPLE ID
106-1
116-1
128-1
AMOUNT SPIKED
1.0
1.0
1.0
D TECH (ppm)
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
HPLC METHOD
8330
0.69
0.73
0.75
AGREEMENT8
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
Y
FP
Y

Y
Y
Y
Y
FP
Y
Y
Y
Y
Y
FP
Y
Y
Y
Y
                                    4050-10
     Revision 0
   January 1995

-------
            TABLE 5 (cont)
RECOVERY OF TNT SPIKED INTO REAL SOILS
SAMPLE ID
106-3
116-3
128-3
AMOUNT SPIKED
3.0
3.0
3.0
D TECH (ppm)
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
0.5 - 1.5
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
0.5 - 1.5
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
0.5 - 1.5
1.5 - 3.0
0.5 - 1.5
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
HPLC METHOD
8330
1.53





2.12
2.07
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
N
Y
FN
Y
Y
Y
Y
Y
Y
Y
Y
Y
               4050-11
  Revision 0
January 1995

-------
                    TABLE 6A
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
SAMPLE ID
61-1
61-10
61-11
61-12
61-13
61-14
61-15
61-16
61-17
61-18
61-19
61-2
61-20
61-21
61-22
61-23
61-24
61-25
61-26
61-27
61-28
61-29
61-3
61-30
61-4
61-5
61-6
61-7
61-8
61-9
TET-1
TET-2
TET-3
TL-1
D TECH RANGE
(ppm)
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
> 1.5
< 0.2
0.5-1.0
< 0.2
< 0.2
1.0-1.5
< 0.2
< 0.2
0.2-0.5
< 0.2
< 0.2
1.0-1.5
< 0.2
> 1.5
0.5 - 1.0
> 1.5
< 0.2
0.5-1.0
0.2-0.5
0.5-1.0
< 0.2
< 0.2
0.2-0.5
METHOD 8330
TNT (ppm)
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
< 0.09
> 3.0
< 0.09
2.44
< 0.09
< 0.09
1.4
< 0.09
< 0.09
0.27
< 0.09
< 0.09
1.3
< 0.09
1.1
1.0
> 3.0
< 0.09
1.0
0.56
< 0.09
< 0.09
< 0.09
0.99
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FN
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
Y
Y
Y
Y
FP
Y
Y
FN
                     4050-12
  Revision 0
January 1995

-------
SAMPLE ID
TL-2
TL-3
TL-4
TL-5
TL-6
TL-7
TL-8
TL-9
D TECH RANGE
(ppm)
r > 1.5
> 1.5
0.2-0.5
> 1.5
0.2-0.5
0.2-0.5
0.5-1.0
0.2-0.5
METHOD 8330
^JNT (ppm)
1.2
> 3.0
0.66
> 3.0
0.66
0.71
1.46
0.92
AGREEMENT3
Y, FN, FP
FP
Y
FN
Y
FN
FN
FN
FN
4050-13
  Revision 0
January 1995

-------
                    TABLE 6B
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample
Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
D TECH
Range (ppm)
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
0.5 - 1.0
0.5 - 1.0
0.5 - 1.0
> 1.5
0.5 - 1.0
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
8330 TNT
(ppm)
5.75
3.32
166
2500
2.72
<2.0
<2.0
140
230
1100
23.5
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
3.23
<2.0
<2.0
4.75
<2.0
<2.0
<2.0
3.64
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
6.39
8330 TNB
(ppm)
< 1.0
< 1.0
< 1.0
18.50
< 1.0
7.02
5.12
12.2
20.2
16.9
11.5
2.95
1.30
1.89
3.94
4.54
4.57
10.5
24.3
81
1.61
2.60
2.97
6.29
< 1.0
5.05
6.62
1.94
8.53
6.77
6.75
17.6
39.2
TNT Equivalent
(ppm)
5.75-6.0
3.32-3.57
166
2504
2.72-2.97
1.76-3.76
1.28-3.28
143
235
1104
26.0
0.74-2.74
0.33-2.33
0.47-2.47
0.99-2.99
1.14-3.14
1.14-3.14
2.63-4.63
9.3
20.3
0.40-2.40
5.40
0.74-2.74
1.57-3.57
<2.25
4.90
1.66-3.66
0.49-2.49
2.13-4.13
1.69-3.69
1.69-3.69
4.40-6.41
16.2
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FN
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
                     4050-14
  Revision 0
January 1995

-------
Sample
Number
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
D TECH
Range (ppm)
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
0.5 - 1.0
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
1.0 - 1.5
> 1.5
> 1.5
> 1.5
8330 TNT
(ppm)
4.20
5.14
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
820
1200
27.6
7.43
4.98
3.32
3.42
4.32
7.57
5.12
<2.0
<2.0
33.5
2.19
7.00
2.84
<2.0
2.23
5.38
2.60
4.43
4.79
2.29
8.84
9.01
29.00
<2.0
8330 TNB
(ppm)
1.39
< 1.0
2.68
7.65
27.70
9.01
30.90
35.70
5.69
24.0
11.9
9.01
9.46
10.4
16.5
28.2
44.8
81.2
1.64
2.27
23.4
8.43
11.0
4.69
5.67
12.8
31.4
13.0
31.1
25.9
18.2
148
< 1.0
6.02
1.30
TNT Equivalent
(ppm)
4.55
5.14-5.39
0.67-2.67
1.91-3.91
6.9-8.9
2.25-4.25
7.7-9.7
8.9-10.9
821
1206
31
9.70
7.40
5.90
7.60
11.4
18.8
25.4
0.41-2.41
0.57-2.57
39.4
4.30
9.75
4.01
1.42-3.42
5.43
13.23
5.85
12.2
11.3
6.8
45.8
9.01
30.50
0.33-2.33
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FN
Y
Y
Y
4050-15
  Revision 0
January 1995

-------
Sample
Number
78
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
95
96
97
98
99
100
101
D TECH
Range (ppm)
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
0.5 - 1.0
0.5 - 1.0
1.0 - 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
1.0 - 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
> 1.5
0.5 - 1.0
> 1.5
> 1.5
> 1.5
> 1.5
< 0.2
> 1.5
> 1.5
8330 TNT
(ppm)
<2.0
<2.0
2.49
<2.0
<2.0
<2.0
<2.0
3.98
5.67
7.05
8.04
1000
2.12
8.83
3.64
3.22
<2.0
<2.0
<2.0
<2.0
<2.0
351
116
4.29
<2.0
2.34
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
4.24
<2.0
8330 TNB
(ppm)
7.50
4.70
30.0
29.1
8.86
30.7
38.1
183
122
< 1.0
< 1.0
7.49
2.99
5.56
3.20
10.6
18.3
17.4
20.4
117
1.96
5.77
39.2
3.92
11.6
9.26
48.7
5.05
12.6
10.7
11.1
3.74
1.88
< 1.0
1.10
TNT Equivalent
(ppm)
1.88-3.88
1.18-3.18
9.99
7.28-9.28
2.22-4.22
7.68-9.68
9.59-11.6
49.7
36.2
7.05-7.3
8.04-8.29
1001
2.87
10.20
4.44
5.87
4.58-6.58
4.43-6.43
5.10-7.10
29.2-31.2
0.49-2.49
352
126
5.27
2.9-4.9
4.66
12.2-14.2
1.26-3.26
3.15-5.15
2.68-4.68
2.78-4.78
0.94-2.94
0.47-2.47
4.24-4.49
0.28-2.28
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FN
FN
FN
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FN
Y
Y
Y
Y
FN
Y
Y
4050-16
  Revision 0
January 1995

-------
Sample
Number
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
through
279
280
through
365
366
through
381
382
through
391
D TECH
Range (ppm)
0.5 - 1.0
1.0 - 1.5
> 1.5
> 1.5
> 1.5
0.5 - 1.0
1.0 - 1.5
0.5 - 1.0
0.5 - 1.0
0.5 - 1.0
1.0 - 1.5
> 1.5
> 1.5
> 1.5
0.2 - 0.5
0.5 - 1.0
0.2 - 0.5
0.5 - 1.0
> 1.5
> 1.5
> 1.5
0.2 - 0.5
< 0.2
< 0.2
0.2 - 0.5
< 0.2
0.2 - 0.5
0.5 - 1.0
1.0 - 1.5
8330 TNT
(ppm)
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
6.35
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
<2.0
8330 TNB
(ppm)
1.28
2.70
10.5
14.1
18.4
6.35
6.66
21.8
5.29
4.49
16.3
28.7
17.7
24.1
< 1.0
2.40
4.70
11.6
56.9
45.6
67.7
2.78
1.61
4.07
3.12
<1.0
< 1.0
< 1.0
< 1.0
TNT Equivalent
(ppm)
0.32-2.32
0.68-2.68
2.63-4.63
3.53-5.53
4.6-6.6
1.59-3.59
1.67-3.67
5.45-7.45
1.32-3.32
1.12-3.12
4.08-6.08
7.18-9.18
4.43-6.43
6.03-8.03
6.35-6.6
0.60-2.6
1.18-3.18
2.9-4.9
14.2-16.2
11.4-13.4
16.9-18.9
0.7-2.7
0.4-2.4
1.02-3.02
0.78-2.78
<2.25
<2.25
<2.25
<2.25
AGREEMENT8
Y, FN, FP
Y
Y
Y
Y
Y
FN
FN
FN
FN
FN
FN
Y
Y
Y
FN
Y
FN
FN
Y
Y
Y
FN
FN
FN
FN
Y
Y
Y
Y
4050-17
  Revision 0
January 1995

-------
Sample
Number
392
through
399
D TECH
Range (ppm)
> 1.5
8330 TNT
(ppm)
<2.0
8330 TNB
(ppm)
< 1.0
TNT Equivalent
(ppm)
<2.25
AGREEMENT3
Y, FN, FP
Y
4050-18
  Revision 0
January 1995

-------
                     TABLE  6C
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
             Third Party Field Trial
Sampl e
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
Oil ution
Factor
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
SDI
Results
<0.5
<0.5
<0.5
0.5-1.5
<0.5
0.5-1.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
0.5-1.5
<0.5
<0.5
0.5-1.5
<0.5
<0.5
<0.5
<0.5
0.5-1.5
<0.5
<0.5
8330
TNT Results
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
8330
TNT+TNB
Results
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
<0.25
AGREEMENT3
Y, FN, FP
Y
Y
Y
FP
Y
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
Y
FP
Y
Y
Y
Y
FP
Y
Y
                     4050-19
  Revision 0
January 1995

-------
Sample
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
Dilution
Factor
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
10
10
1
SDI
Results
<0.5
0.5-1.5
<0.5
0.5-1.5
<0.5
0.5-1.5
0.5-1.5
0.5-1.5
<0.5
<0.5
3.0-4.0
<0.5
<0.5
<0.5
0.5-1.5
1.5-3.0
<0.5
0.5-1.5
0.5-1.5
0.5-1.5
0.5-1.5
<0.5
0.5-1.5
<0.5
0.5-1.5
5-15
40-50
0.5-1.5
8330
TNT Results
<0.15
<0.15
<0.15
0.15-0.99
<0.15
<0.15
<0.15
0.15-0.99
<0.15
<0.15
0.15-0.99
<0.15
<0.15
<0.15
<0.15
0.15-0.99
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
<0.15
1.4
35
<0.15
8330
TNT+TNB
Results
<0.25
<0.25
<0.25
0.15-0.99
<0.25
<0.25
<0.25
0.15-0.99
<0.25
<0.25
0.25-2.0
<0.25
<0.25
<0.25
<0.25
0.15-0.99
<0.25
0.15-0.99
<0.25
<0.25
1.3
<0.25
<0.25
<0.25
<0.25
3.2
41.67
<0.15
AGREEMENT3
Y, FN, FP
Y
FP
Y
Y
Y
FP
FP
Y
Y
Y
FP
Y
Y
Y
FP
FP
Y
Y
FP
FP
Y
Y
FP
Y
FP
Y
Y
FP
4050-20
  Revision 0
January 1995

-------
Sample
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
Dilution
Factor
1
1
1
1
1
10
1
10
10
100
1000
10000
1000
10
100
10
10
10
10
10
10
10
10
100
10
10
10
10
SDI
Results
0.5-1.5
1.5
0.5-1.5
3.0-4.0
0.5-1.5
15-30
0.5-1.5
4-40
5-15
400-500
4000-5000
15000
15000
5-15
400-500
15-30
5-15
40
5-15
5
4-30
5-15
5-15
300-400
5-15
5-15
5-15
15-30
8330
TNT Results
0.15-0.99
0.15-0.99
<0.15
0.15-0.99
<0.15
22
-
2.1
2
360
6300
4000
530
2.8
460
4.2
1.0
5.1
1.9
1.6
2.2
1.7
2.2
180
3.1
2.8
2.5
3.2
8330
TNT+TNB
Results
0.15-0.99
0.15-0.99
<0.15
0.15-0.99
<0.15
22.48
<0.15
32
3.1
364
6327
4027
547
3.375
477
6.73
1.57
34.5
4
2.7
4.3
2
3.95
192.19
4.61
5.26
5.26
4.5
AGREEMENT3
Y, FN, FP
Y
Y
FP
FP
FP
Y
FP
Y
Y
Y
Y
FP
FP
Y
Y
FP
FP
Y
Y
Y
Y
FP
Y
Y
Y
Y
Y
FP
4050-21
  Revision 0
January 1995

-------
Sample
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
Dilution
Factor
10
10
10
10
10
10
100
1
10000
10
10000
1000
1
10000
10000
10
10
100
SDI
Results
40-50
15-30
15-30
15-30
5-15
5-15
150-300
4-5
15000
40-50
15000-30000
500-1500
3.0
40000-50000
4000-5000
15-30
15-30
50-150
8330
TNT Results
1
3.8
36
3.6
2.6
3.2
78
18000
11000
36
11000
88
9.6
15000
2200
3.6
6.4
26
8330
TNT+TNB
Results
23
18.5
52.5
8.66
19.16
3.84
82
18050
11052.9
42.4
11052.9
107
10.17
15050
2220
3.9
6.7
28.76
AGREEMENT3
Y, FN, FP
Y
Y
FN
Y
Y
Y
Y
FN
Y
Y
Y
FP
FN
FP
Y
FP
FP
Y
4050-22
  Revision 0
January 1995

-------
                                  METHOD 4051

 HEXAHYDRO-1.3.5-TRINITRO-1,3.5-TRIAZINE (RDX) IN SOIL & WATER BY IMMUNOASSAY
1.0   SCOPE AND APPLICATION

      1.1   Method  4051  is  a  procedure  for  screening waters  and  soils  to
determine when hexahydro-l,3,5-trinitro-l,3,5-triazine  (RDX, CAS No.  121-82-4)
is  present  at concentrations  above 5  p,g/L in  water  and  0.5 mg/kg  in soil.
Method 4051 provides an estimate of the concentration of RDX by comparison with
a reference.

      1.2   Using the test kit from which this method was developed,  96% of water
samples containing 2.5 ppb or less of RDX will produce a negative  result and 99+%
of waters containing 10 ppb or more will produce a positive  result.   In  addition
99+% of soil samples containing 0.25 ppm or  less  of RDX  will produce a  negative
result and  99+% of  soil  samples containing 1.0  ppm will   produce  a   positive
result.

      1.3   In  cases  where  the  exact  concentration   of RDX  is   required,
quantitative techniques (i.e., Method 8330)  should be used.

      1.4   This method  is restricted  to use  by or under  the  supervision  of
trained  analysts.    Each  analyst  must demonstrate  the  ability  to   generate
acceptable results with this method.
2.0   SUMMARY OF METHOD

      2.1   Test  kits  are  commercially  available   for  this  method.    The
manufacturer's  directions should  be  followed.    In  general,  the  method  is
performed using a diluted water sample or an  extract of a soil sample.  Samples
and an  enzyme  conjugate reagent are  added to immobilized RDX  antibody.   The
enzyme-RDX  conjugate "competes" with RDX present in the sample for binding to
immobilized RDX antibody.  The enzyme-RDX conjugate bound to the antibody then
catalyzes a colorless substrate to a colored product.   The test is interpreted
by comparing  the color  produced  by  a  sample to the  response produced  by  a
reference reaction.
3.0   INTERFERENCES

      3.1   Chemically similar compounds and compounds which might be expected
to be found in conjunction with RDX contamination were tested to determine the
concentration required to produce a positive test result.

            3.1.1  Table  1 provides the concentrations of compounds tested  with
      the D TECH test  kit that are required to elicit  a  positive response at the
                                    4051-1
  Revision 0
January 1995

-------
      MDL,  as  well  as  the  concentration  required  to  yield 50%  inhibition
      compared to the standard curve.


4.0   APPARATUS AND MATERIALS

      4.1   Immunoassay test kit:  D TECH™ RDX (Strategic  Diagnostics Inc.),  or
equivalent.  Each  commercially  available test kit will  supply or  specify the
apparatus and materials necessary for successful  completion of the test.


5.0   REAGENTS

      5.1   Each commercially  available test kit  will  supply or  specify the
reagents necessary for successful completion of the test.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HAULING

      6.1   See the  introductory  material  to this chapter,  Organic  Analytes,
Section 4.1.

      6.2   Water and soil  samples may be contaminated, and should therefore be
considered hazardous and handled accordingly.


7.0   PROCEDURE

      7.1   Follow the manufacturer's instructions  for  the test kit being used.
Those test kits used must meet or exceed the performance indicated  in Tables 3-6.


8.0   QUALITY CONTROL

      8.1   Follow the manufacturer's instructions for the test kit being used
for quality control  procedures  specific to the test kit  used.   Additionally,
guidance provided in Chapter One should  be followed.

      8.2   Use  of  replicate  analyses,  particularly when   results  indicate
concentrations near  the action  level,   is  recommended to  refine  information
gathered with the kit.

      8.3   Do not use test kits past their expiration date.

      8.4   Do not use tubes or reagents designated  for use with other test kits.
Do not mix reagents from one kit lot with a different kit lot.

      8.5   Use the  test kits within their  specified  storage temperature and
operating temperature limits.
                                    4051-2                          Revision 0
                                                                  January 1995

-------
      8.6   Method 4051  is intended for field or laboratory use.  The appropriate
level of quality  assurance  should  accompany the  application  of this method to
document data quality.


9.0   METHOD PERFORMANCE

      9.1   Table  1  provides data on  the minimum concentrations  of possible
interferants and co-contaminants required to elicit a positive response in the
test kits evaluated.

      9.2   Twenty six soil  samples,  known to not be contaminated with RDX, were
extracted and analyzed using the DTECH RDX kit to determine the extent of soil
matrix effects on the performance of  the  test  kit.  The results are provided in
Table 2A,  and  show that false  positive  results  are  not  attributable  to soil
components.  Table 2B presents similar data generated from the  analysis of thirty
uncontaminated water samples.

      9.3   Thirty  water  samples  and  thirty  soil  samples,   known  to  not  be
contaminated with RDX, were each spiked with RDX at one-half and two times the
MDL (0.25 and 1.0 ppm respectively).  These samples were analyzed with the DTECH
RDX test kit to determine the error rate of the assay.   The  results are presented
in Tables 3A and 3B.

      9.4   Ten different  soil  types, all known to not  be contaminated with RDX,
were spiked with  RDX.  The spiked soil samples  were each analyzed six times with
the DTECH kit to determine the extraction  efficiency of the method.  The data are
presented in Table 4.

      9.5   Table 5A presents the results of analysis of three soils spiked at
approximately 0.4, 1 and 3  ppm RDX.   Each  sample was analyzed using Method 8330
and in triplicate using  the  DTECH kit.  Table 5B presents similar data generated
using water samples spiked  at 10, 20 and 40 ppb of RDX.

      9.6   Tables  6A  through 6D  present the results  of four  field  trials.
Freshly collected  (Table  6A,  6B and  6D)  and  archived (6C)  soil  samples,  and
samples of water collected  from  monitoring wells  (Table 6B), were analyzed by
commercial laboratories using Method 8330 and  the DTech  test kit.   The Tables
provide results for both analyses, and evaluate the agreement between the two.


10.0  REFERENCES

1.    D TECH™ TNT Users Guide , SDI/Em Sciences.

2.    Haas, R.J.,  and B.P.  Simmons,  "Measurement  of  Trinitrotoluene (TNT)  and
      Hexahydro-l,3,5-trinitro-l,3,5-triazine    (RDX)    in   Soil   by   Enzyme
      Immunoassay and High Performance Liquid Chromatography (EPA Method 8330)",
      California Environmental  Protection Agency, Department of Toxic Substances
      Control,  Hazardous Materials Laboratory, March,  1995.


                                    4051-3                          Revision 0
                                                                  January 1995

-------
                                 TABLE  1
                 CROSS  REACTANTS  -  D  TECH™  RDX test  kit
SAMPLE
RDX d
HMX d
TNT (trinitrotoluene)
Tetryl d
TNB (trinitrobenzene)
2-amino-4,6-dinitrotoluene
4-amino-2,6-dinitrotoluene
2,4-dinitrotoluene
2,6-dinitrotoluene
1,3-dinitrobenzene
nitrobenzene
2-nitrotoluene
3-nitrotoluene
4-nitrotoluene
nitroglycerine
pent aery thritoltetranitrate
MDLa
(Ppb)
5
150
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
ICgOb
(ppb)
25
800
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
> 500
% CROSS
REACTIVITY0
100
3
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
< 1
The following compounds were not detected at or above 500 ppm (lOOx the method
MDL for RDX):
Atrazine Benzo(a)pyrene Benzo(b)fl uoranthene Benzene
Aroclor 1254 Acenaphthene Dibenz(ah)anthracene Chrysene
Acetone Acenaphthalene Fluoranthene Fluorene
Toluene 1,2-Benzanthracene Benzo(k)fluoranthene Pyrene
Ethylbenzene Indeno(123-cd)pyrene Benzo(ghi )perylene Xylene
Naphthalene Methanol Phenanthrene
The Method Detection Limit (MDL) is defined as the lowest concentration of
compound that yields a positive test result.
The IC50 is  defined as  the concentration  of compound  required  to  produce a
test response equivalent to 50% of the maximum response.
% Cross Reactivity is determined by dividing the equivalent RDX
concentration by the actual compound concentration at IC50.
RDX = hexahydro-l,3,5-trinitro-l,3,5-triazine
HMX = octahydro-l,3,5,7-tetranitro-l,3,5,7-tetrazocine
Tetryl = methyl-2,4,6-trinitrophenylnitramine
                                 4051-4
  Revision 0
January 1995

-------
     TABLE 2A
SOIL MATRIX EFFECTS
Soil ID #
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
Soil Type
Low OM Clay Loam
Sassafras Sandy Loam
Cecil Sandy Clay Loam
Davidson Clay Loam
Shontik-Casa Grande Clay
Trix Sandy Clay Loam
Trix-Casa Grande Clay Loam
Yolo Loam
Capay Silty Clay
Sycamore Silt Loam
Dennis Silt Loam
Luray Silty Clay Loam
Wooster Silt Loam
Vienna Loam
Opal Clay
Raulb Silt Loam
Rockfield Silt Loam
Cisne Silt Loam
Muscatine Silt Loam
Avonburg
Matapeake Silt Loam
Evesboro Low OM Sand
Selbyville High OM Sand
Casa Grande Clay Loam
Grundy Silty Clay Loam
Drummer Silty Clay
Non-Soil Control
State
DE
DE
GA
GA
AZ
AZ
AZ
CA
CA
CA
KA
OH
OH
SD
SD
IN
IN
IL
IL

DE
DE
DE
AZ
KA
IL
-
D TECH Result
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
      4051-5
  Revision 0
January 1995

-------
      TABLE 2B
WATER MATRIX EFFECTS
Water ID
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
Location
Adamsville, RI
Buttermilk Falls, PA
Hudson River, PA
Germantown, PA
Houston, TX (1)
Houston, TX (2)
Ontario, CA
Pacific Ocean, CA
S. Darthmouth, MA
Newark, DE (1)
U.S. Army Waterways
U.S. Army WES
U.S. Army WES
U.S. Army WES
U.S. Army WES
U.S. Army WES
U.S. Army WES
U.S. Army WES
Georgetown, DE
Newark, DE (2)
Burlington, IA
Burlington, IA
Lake St. Germain, Canada
Milliston, WI
Moorhead, MN
McKenzie Co., ND
Wolcott, IN
Newark, DE (3)
Smith Island, MD
Adrian, GA
DI Control
D TECH Result (ppb)
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
< 5.0
       4051-6
  Revision 0
January 1995

-------
                                   TABLE 3A
            False Negative and False Positive Rates, Soil Matrix3
Spike Concentration
0.25 ppm
1.0 ppm
False Positive Rate
0%
-
False Negative Rate
-
0%
3 Thirty negative soils were spiked with RDX at one-half and two times the MDL
(0.25 and 1.0 ppm respectively).  These samples were analyzed with the DTECH
RDX test kit to determine the error rate of the assay.
                                   TABLE 3B
            False Negative and False  Positive Rates,  Water Matrix3
Spike Concentration
0.25 ppm
1.0 ppm
False Positive Rate
3.3%
-
False Negative Rate
_
0%
a Thirty negative water samples were spiked with RDX at one-half and two times
the MDL (0.25 and 1.0 ppm respectively).  These samples were analyzed in
triplicate with the DTECH RDX test kit to determine the error rate of the
assay.
                                    4051-7
  Revision 0
January 1995

-------
                         TABLE 4
DETERMINATION OF EXTRACTION EFFICIENCY FROM SOIL SAMPLES3
Soil ID : Spike
(ppm)
101:1
106:1
108:1
109:1
110:1
116:1
117:1
123:1
126:1
128:1
Non-Soil
Average
101:6
106:6
108:6
109:6
110:6
116:6
117:6
123:6
126:6
128:6
Non-Soil
Average
Mean RDX
Concentration
(ppm)
0.53
0.88
0.86
0.66
0.70
0.96
0.92
1.00
1.03
1.02
1.05
0.86
4.9?
6.15
5.69
6.11
6.12
6.26
5.71
6.05
6.82
6.02
6.02
5.98
Standard
Deviation
0.19
0.13
0.23
0.22
0.14
0.12
0.42
0.45
0.25
0.18
0.13
0.23
0.54
0.84
1.09
0.93
0.46
1.21
0.72
0.8
0.33
0.62
0.83
0.75
Coefficient
of
Variation
(%)
35
15
26
34
19
13
46
45
24
18
12
27
11
14
19
15
8
19
13
13
5
10
14
13
Recovery
(a/ \
\'°)
53
88
86
66
70
96
92
100
103
102
105
86
8?
103
95
102
102
104
95
101
114
100
100
100
                          4051-8
  Revision 0
January 1995

-------
               TABLE  5A
RECOVERY OF RDX SPIKED INTO REAL SOILS.
Soil ID
106
116
128
Spike
Concentration
(ppm)
0.4


1.0


3.0


0.4


1.0


3.0


0.4


1.0


3.0


Method 8330
(ppm)
0.32


0.83


1.79


0.29


0.66


0.61


0.31(0.25)


0.73(0.73)


0.75(2.27)


D TECH
(ppm)
< 0.5
< 0.5
< 0.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
> 2.0
> 2.0
> 9 n
< 0.5
< 0.5
< 0.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
> 2.0
< 2.0
> ? n
< 0.5
< 0.5
< 0.5
< 0.5
0.5 - 1.5
0.5 - 1.5
> 2.0
< 2.0
< ? n
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
FP
FP
FP
Y
Y
Y
Y
Y
Y
FP
FP
FP
Y
Y
Y
FN
Y
Y
Y
Y
Y
                4051-9
  Revision 0
January 1995

-------
TABLE 5B
RECOVERY OF RDX SPIKED INTO WATERS
Sample ID
1











7











Spike
Concentration
(ppb)
10


20


40


10


10


20


40


20


Method 8330
(ppb)
11.1


18.0


35.7


9.0


9.?


19.4


36.5


17.1


D TECH
(ppb)
5 - 15
5 - 15
5 - 15
15 - 30
15 - 30
15 - 30
> 45
> 45
> 45
5 - 15
5 - 15
5 - 15
5 - 15
5 - 15
5 - 15
15 - 30
15 - 30
15 - 30
> 45
> 45
> 45
15 - 30
15 - 30
15 - 30
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
FP
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
FP
FP
Y
Y
Y
4051-10
  Revision 0
January 1995

-------
TABLE 5B
RECOVERY OF RDX SPIKED INTO WATERS
Sample ID
3











Spike
Concentration
(ppb)
10


20


40


40


Method 8330
(ppb)
9.7


18.2


35.8


31.8


D TECH
(Ppb)
5 - 15
5 - 15
5 - 15
15 - 30
15 - 30
15 - 30
> 45
> 45
30 - 45
> 45
30 - 45
30 - 45
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
FP
FP
Y
FP
Y
Y
4051-11
  Revision 0
January 1995

-------
TABLE 6A
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample ID
S4
S12
S14
S15
S19
S20
S21
Tl-2
T2-4
T6-1
T3-5
T12-3
T12-6
T20-3
T21-10
T22-4
T22-5
T22-6
T28-3
T28-4
T28-5
T28-6
T28-7
T28-8
T28-9
Method 8330
(ppm)
< 0.2
< 0.2
1.72
< 0.2
2.12
1.61
0.32
0.21
1.41
2.62
2.00
< 0.2
1.00
< 0.2
1.89
< 0.2
0.83
0.99
3.73
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
< 0.2
D TECH
(ppm)
< 0.5
< 0.5
1.5 - 2.0
< 0.5
1.5 - 3.0
1.5 - 3.0
< 0.5
< 0.5
1.5 - 2.0
> 3.0
0.5 - 1.5
< 0.5
0.5 - 1.5
< 0.5
1.5 - 2.0
< 0.5
0.5 - 1.5
0.5 - 1.5
> 3.0
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
FP
FP
FN
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
4051-12
  Revision 0
January 1995

-------
TABLE 6A
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample ID
T28-10
T28-11
T28-12
T28-13
T31-4
T12-5
Method 8330
(ppm)
0.28
1.51
1.3
0.6
1.22
0.26
D TECH
(ppm)
< 0.5
1.5 - 3.0
1.5 - 3.0
0.5 - 1.5
1.5 - 2.0
< 0.5
AGREEMENT3
Y, FN, FP
Y
Y
FP
Y
FP
Y
4051-13
  Revision 0
January 1995

-------
TABLE 6B
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330

Sample
ID
1
3
13
15
16
23
24
25
26
31
33
34
35
37
38
43
44
47
48
58
59
64
67
68
75
84
85
87
94
96
97
Method 8330
(ppm)
4.00
19.0
1.30
1.80
3.40
0.48
0.68
0.68
0.75
0.13
0.74
0.48
1.30
5.50
0.55
1.30
40.0
2.30
0.36
0.79
0.80
2.20
10.9
3.40
3.90
17.6
70.3
101
1.60
0.20
5.40
Replicate 1
D TECH
(ppm)
> 3.0
> 6.0
0.5 - 1.5
1.5 - 3.0
> 3.0
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
> 6.0
0.5 - 1.5
1.5 - 3.0
> 6.0
> 3.0
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
1.5 - 3.0
> 6.0
1.5 - 3.0
> 3.0
> 6.0
> 6.0
> 6.0
1.5 - 3.0
< 0.5
> 3.0
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
Y
FP
FP
Y
FP
Y
FP
FP
Y
FP
Y
Y
FN
Y
Y
Y
Y
Y
Y
Y
Replicate 2
D TECH
(ppm)
> 3.0
> 6.0
0.5 - 1.5
1.5 - 3.0
> 3.0
0.5 - 1.5
< 0.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
0.5 - 1.5
1.5 - 3.0
> 3.0
0.5 - 1.5
1.5 - 3.0
> 6.0
> 3.0
< 0.5
0.5 - 1.5
1.5 - 3.0
1.5 - 3.0
> 6.0
1.5 - 3.0
> 3.0
> 6.0
> 6.0
> 6.0
1.5 - 3.0
< 0.5
> 3.0
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
Y
FN
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
FP
Y
Y
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
4051-14
  Revision 0
January 1995

-------
TABLE 6B
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330

Sample
ID
98
99
105
111
113
115
119
Method 8330
(ppm)
< 0.05
< 0.05
130
< 1.0
< 1.0
3.00
36.0
Replicate 1
D TECH
(ppm)
< 0.5
< 0.5
> 60
> 3.0
< 5.0
< 5.0
> 30
AGREEMENT3
Y, FN, FP
Y
Y
Y
FP
Y
Y
Y
Replicate 2
D TECH
(ppm)
< 0.5
0.5 - 1.5
> 60
< 5.0
< 0.5
< 0,5
15 - 30
AGREEMENT3
Y, FN, FP
Y
FP
Y
Y
FN
FN
FN
4051-15
  Revision 0
January 1995

-------
TABLE 6C
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample ID
1
2
3
4
5
6
7
8
9
10
19
20
11
12
13
14
15
16
17
18
21
22
23
24
25
METHOD 8330
(ppm)
17
34
48
160
650
41
360
840
69
85
17
19
4.3
1.9
4.9
27
1.2
1.0
0.82
0.78
0.67
0.94
< 0.4
< 0.4
< 0.4
D TECH
(ppm)
15 - 30
15 - 30
> 30
60 - 120
150 - 300
> 30
50 - 150
> 600
> 60
30 - 60
> 6.0
> 6.0
> 3.0
> 3.0
> 3.0
1.5 - 3.0
1.5 - 3.0
1.5 - 3.0
0.5 - 1.5
0.5 - 1.5
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
AGREEMENT3
Y, FN, FP
Y
FN
Y
FN
FN
Y
FN
Y
Y
FN
Y
Y
Y
FP
Y
Y
FP
FP
Y
Y
FN
FN
Y
Y
Y
4051-16
  Revision 0
January 1995

-------
TABLE 6C
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample ID
26
27
28
29
30
METHOD 8330
(ppm)
< 0.4
< 0.4
< 0.4
< 0.4
< 0.4
D TECH
(ppm)
< 0.5
< 0.5
< 0.5
< 0.5
< 0.5
AGREEMENT3
Y, FN, FP
Y
Y
Y
Y
Y
4051-17
  Revision 0
January 1995

-------
TABLE 6D
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
SDI
Dilution
Factor
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
SDI
Results
<0.5
<0.5
<0.5
<0.5
0.5-1.5
<0.5
<0.5
0.5-1.5
<0.5
<0.5
0.5-1.5
0.5-1.5
<0.5
<0.5
<0.5'
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
0.5-1.5
<0.5
<0.5
8330
Results
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
AGREEMENT8
Y, FN, FP
Y
Y
Y
Y
FP
Y
Y
FP
Y
Y
FP
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
FP
Y
Y
4051-18
  Revision 0
January 1995

-------
TABLE 6D
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
SDI
Dilution
Factor
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
SDI
Results
0.5-1.5
<0.5
<0.5
1.5-3.5
<0.5
0.5-1.5
<0.5
0.5-1.5
0.5-1.5
<0.5
1.5-3.0
<0.5
0.5-1.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
1.5-3.0
<0.5
0.5-1.5
<0.5
0.5-1.5
<0.5
0.5-1.5
8330
Results
<0.17
<0.17
<0.17
0.17-0.99
<0.17
<0.17
<0.17
0.17-0.99
0.17-0.99
<0.17
1.2
<0.17
<0.17
<0.17
<0.17
3.8
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
0.17-0.99
AGREEMENT8
Y, FN, FP
FP
Y
Y
FP
Y
FP
Y
Y
Y
Y
FP
Y
FP
Y
Y
FN
Y
Y
Y
FP
Y
FP
Y
FP
Y
Y
4051-19
  Revision 0
January 1995

-------
TABLE 6D
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
SDI
Dilution
Factor
100
1
1
1
1
1
1
1
1
100
10
1
1
1
100
1
100
1
1
100
10
10
10
1
1
10
SDI
Results
50-150
3.0-4.5
3.0-4.5
<0.5
0.5-1.5
<0.5
<0.5
1.5-3.0
0.5-1.5
150-300
15-30
1.5-3.0
1.5-3.0
3.0-4.5
50-150
0.5-1.5
50-150
0.5-1.5
1.5-3.0
150-300
45-60
>60
30-45
1.5-3.0
4.5-6.0
>60
8330
Results
100
<0.17
<0.17
<0.17
<0.17
<0.17
<0.17
1.1
<0.17
290
46
4.8
0.17-0.99
12
150
2.6
140
7.8
3.2
340
55
67
63
2.4
6.4
73
AGREEMENT8
Y, FN, FP
Y
FP
FP
Y
FP
Y
Y
FP
FP
Y
FN
FN
FP
FN
Y
FN
Y
FN
FN
FN
Y
Y
FN
Y
FP
Y
4051-20
  Revision 0
January 1995

-------
TABLE 6D
COMPARISON OF DTECH SOIL RESULTS WITH METHOD 8330
Sample
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
SDI
Dilution
Factor
10
1
1
1
10
1
100
10
100
1
1
1
1
1
1
1
1
1
1
1
1
1
SDI
Results
15-30
0.5-1.5
3.0-4.5
1.5-3.0
>60
>6
50-150
30-45
50-150
0.5-1.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
<0.5
8330
Results
14
2.1
2.4
2
94
23
150
34
150
1.2
0.17-0.99
<15
<15
<2
<15
<5
<0.17
<15
<5
<0.17
<0.17
<0.17
AGREEMENT8
Y, FN, FP
FP
FN
FP
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
Y
4051-21
  Revision 0
January 1995

-------
4.5  MISCELLANEOUS SCREENING METHODS

         The following methods are included in this section:

         Method 3810:      Headspace
         Method 3820:      Hexadecane Extraction and Screening of Purgeable
                           Organics
         Method 8275A:     Semivolatile Organic Compounds (PAHs and PCBs) in
                           Soils/Sludges  and  Solid Wastes  Using  Thermal
                           Extraction/Gas  Chromatography/Mass  Spectrometry
                           (TE/GC/MS) for Screening
         Method 8515:      Colorimetric Screening Method  for Trinitrotoluene
                           (TNT) in Soil
         Method 9078:      Screening   Test   Method  for   Polychlorinated
                           Biphenyls in Soil
         Method 9079:      Screening   Test   Method  for   Polychlorinated
                           Biphenyls in Transformer Oil
                                 FOUR -  16                       Revision 3
                                                               January 1995

-------
                                 METHOD 8275A

                SEMIVOLATILE ORGANIC COMPOUNDS (PAHs and PCBs)
                    IN  SOILS/SLUDGES  AND  SOLID WASTES USING
      THERMAL EXTRACTION/GAS CHROMAT06RAPHY/MASS SPECTROMETRY (TE/GC/MS)
1.0  SCOPE AND APPLICATION

     1.1  Method 8275 is a thermal  extraction  capillary GC/MS procedure for the
rapid quantitative  determination of targeted PCBs and PAHs  in soils, sludges and
solid wastes.  The  following analytes can be determined by this method:
     Compound
                                  8275A - 1
CAS No.'
Acenaphthene
Acenaphthylene
Anthracene
Benz[a]anthracene
Benzo[a]pyrene
Benzo[b]fluoranthene
Benzo[g,h,i]perylene
Benzo[k]fluoranthene
4-Bromophenyl phenyl ether
1-Chloronaphthalene
Chrysene
Dibenzofuran
Dibenz[ a, h] anthracene
Dibenzothiophene
Fluoranthene
Fluorene
Hexachlorobenzene
Indeno(l,2,3-cd)pyrene
Naphthalene
Phenanthrene
Pyrene
1 ,2,4-Trichlorobenzene
2-Chlorobiphenyl
3,3'-Dichlorobiphenyl
2,2' ,5-Trichlorobiphenyl
2,3' ,5-Trichlorobiphenyl
2,4' ,5-Trichlorobiphenyl
2,2' ,5,5'-Tetrachlorobiphenyl
2,2',4,5'-Tetrachlorobiphenyl
2,2',3,5'-Tetrachlorobiphenyl
2,3' ,4,4'-Tetrach1orobiphenyl
2,2',4,5,5'-Pentachlorobiphenyl
2,3' ,4,4' ,5-Pentachlorobiphenyl
2,2' ,3,4,4',5'-Hexachlorobiphenyl
2,2',3,4',5,5',6-Heptachlorobiphenyl
83-32-9
208-96-8
120-12-7
56-55-3
50-32-8
205-99-2
191-24-2
207-08-9
101-55-3
90-13-1
218-01-9
132-64-9
53-70-3
132-65-0
206-44-0
86-73-7
118-74-1
193-39-5
91-20-3
85-01-8
129-00-0
120-82-1
2051-60-7
2050-67-1
37680-65-2
3844-81-4
16606-02-3
35693-99-3
41464-40-8
41464-39-5
32598-10-0
37680-73-2
31508-00-6
35065-28-2
52663-68-0
            Revision  1
          January  1995

-------
      Compound                                           CAS No.'
2,2',3,3/,4,4'-Hexachlorobiphenyl
2,2',3,4,4',5,5'-Heptachlorobiphenyl
2,2',3,3',4,4',5-Heptachlorobiphenyl
2,2',3,3',4,4',5,5'-Octachlorobiphenyl
2,2',3,3',4,4',5,5',6-Nonachlorobiphenyl
2,2',3,3',4,4',5,5',6,6'-Decachlorobiphenyl
a Chemical Abstract Services Registry Number.
38380-07-3
35065-29-3
35065-30-6
35694-08-7
40186-72-9
2051-24-3

      1.2   The  estimated  quantitation limit  (EQL) of Method 8275 for individual
PAH compounds is 1.0 mg/kg  (dry weight)  (0.2 mg/kg for  individual PCB congeners)
for  soil/sediment  samples  and  75 mg/kg  for wet  sludges/other  solid  wastes
(depending on  water  and solvent  content).   However,  this  can be  lowered  by
adjusting the range of the  calibration curve or introducing larger sample sizes
if sample  interferences  are  not  a  factor.   Detection  limits  achievable during
method development ranged  from  0.01  to  0.5 mg/kg for  compounds  in  the target
analyte list in Section 1.1  (dry samples).

      1.3   This method is  restricted to use  by or  under the  supervision  of
analysts  experienced  in  the  operation   of  a  gas  chromatograph  and  mass
spectrometer and  skilled in the  interpretation of  mass spectral data.   Each
analyst must demonstrate  the  ability to maintain  control  and generate acceptable
results with this method.
2.0   SUMMARY  OF  METHOD

      2.1  A   portion  of  sample   (0.003-0.250  g,  depending  on  the  expected
concentration) is weighed into a sample crucible.

      2.2  The crucible is placed  in a thermal extraction chamber and then heated
to 340°C where it is held for 3 minutes.

      2.3  Thermally-extracted  compounds are swept  into a GC equipped  with  a
split/splitless injection port  (split ratio set at -35:1 for  a low concentration
sample or -400:1 for a high concentration sample) and then concentrated on the
head of GC column.  Thermal desorption lasts 13 minutes.

      2.4  The temperature program of the GC oven  is  adjusted to  the specific
temperature conditions  required  to  elute  the  target  analytes.   The  target
analytes are  swept  into  a mass spectrometer for  qualitative  and  quantitative
determination.
                                   8275A  -  2                         Revision 1
                                                                  January 1995

-------
3.0   INTERFERENCES

      3.1   Raw  GC/MS  data from all blanks,  samples,  calibration  standards and
internal standards must be evaluated for interferences.

      3.2   Whenever a  heavily concentrated  sample is  encountered  the GC column
can become  over  loaded and  an ON-LINE  bakeout (Section 7.2.1)  followed  by a
method blank is necessary.

      3.3   A maintenance bakeout (Section 7.5)  is  performed whenever the ON-LINE
bakeout and subsequent blank analyses do not eliminate system contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Thermal  extraction/gas  chromatograph/mass  spectrometer  (TE/GC/MS)
system

           4.1.1   Mass spectrometer  -  Capable of scanning from  35  to  500 amu
      every  1  sec or  less,  using  70 volts (nominal)  electron energy  in the
      electron impact ionization mode.  The  mass spectrometer must be capable of
      producing a  mass  spectrum  for decafluorotriphenylphosphine  (DFTPP) which
      meets  the criteria of Method  8270.

           4.1.2   Data system - A  computer  interfaced to the mass spectrometer
      should allow the continuous  acquisition  and storage  on  machine-readable
      media  of  all  mass   spectra  obtained  throughout  the  duration  of  the
      chromatographic program.  The computer must have software that can search
      the GC/MS data  file for  ions of  a specific mass  and  that can plot such ion
      abundances versus time  or scan number.  This type of plot is defined as a
      reconstructed ion chromatogram (RIC).  Software must also be available that
      allows integrating  the  abundances of  the RIC  between specified  time  or
      scan-number  1imits.

           4.1.3   GC/MS interface - Any GC-to-MS interface that gives acceptable
      calibration  points in the concentration range of  interest may be used.

           4.1.4   Gas   chromatograph  -   Must  be  equipped   with   a   heated
      split/splitless capillary injection port, column  oven,  cryogenic  cooling
      (optional).  The  oven temperature should  be  controllable  from  ambient  to
      450°C, and  have  programmable oven heating  controls capable of rates  of
      rc/min to 70°C/min.

           4.1.5   Recommended capillary column -  A fused  silica coated with (5%
      phenyl)-methylpolysiloxane phase; 25-50 meter length x  (0.25 to 0.32 mm)
      I.D. with 0.1 to  1.0  micron film thickness (OV-5 or equivalent),  depending
      on analyte volatility and separation requirements.

          4.1.6   Thermal  extraction unit  - The  TE unit  should be  constructed
      such that the sample and any compounds extracted are permitted  to  contact
      only heated fused quartz surfaces during the  extraction and transfer to the
      GC injection port.   It  is also  imperative  that all zones  in the  sample
      transfer path be  kept at a  minimum  of 315°C.  The  unit must also  have a
      bakeout capability of at least 650°C in the thermal extraction chamber and

                                   8275A - 3                         Revision 1
                                                                 January  1995

-------
     450°C  in the interface zone.  It should  also be noted that all components,
     crucibles,  spatulas and tools  that come in  contact  with the  sample  be
     constructed of  fused quartz to  permit total oxidation of  any residues.

     4.2  Fused quartz  sample spatula.

     4.3  Muffle furnace tray -  for  holding  the crucibles while cleaning.

     4.4  Stainless  steel forceps  for sample crucible handling.

     4.5  Petri dishes - for sample crucibles; one for clean storage and one for
dirty storage.

     4.6  Sample staging disk.

     4.7  Porous fused  quartz sample crucibles.

     4.8  Porous fused  quartz sample crucible lids.

     4.9  Muffle furnace - for cleaning  sample crucibles, capable of heating to
800°C.

     4.10 Cooling rack/pad  - high  temperature, ceramic or quartz.

     4.11   Analytical balance - minimum  2 gram capacity, 0.01  mg sensitivity.

     4.12   Mortar and pestle.

     4.13   100- and  60-mesh  sieves.

     4.14   Sample vials  - glass,  with Teflon®-lined caps.


5.0  REAGENTS

     5.1  Organic-free  reagent water -  All references to  water in this method
refer to organic-free reagent water,  as  defined  in  Chapter One.

     5.2  Stock  standard solutions  (1000  mg/L)  -  Standard  solutions can  be
prepared from pure standard materials or purchased  as certified solutions.

          5.2.1   Prepare stock standard solutions  by accurately weighing about
     0.0100 g of  pure material.   Dissolve  the  material in  pesticide  quality
     methylene chloride or other  suitable solvent (some PAHs may require initial
     dissolution in  small volumes of  toluene  or carbon disulfide) and dilute to
     volume in  a  10-mL  volumetric flask.  Larger  volumes can be used  at the
     convenience of  the  analyst.  When compound purity is assayed to be 96% or
     greater,  the weight  may be  used  without  correction  to calculate  the
     concentration of the stock standard. Commercially-prepared stock standards
     may be used at  any concentration if they are certified by the manufacturer
     or by  an independent source.

          5.2.2   Transfer  the  stock  standard  solutions into  bottles  with
     Teflon®-!ined screw-caps. Store at  -10°C to -20°C or less  and protect from

                                   8275A  - 4                         Revision 1
                                                                  January 1995

-------
      light.  Stock standard solutions should be checked frequently  for signs of
      degradation or evaporation, especially just prior to preparing calibration
      standards  from  them.

           5.2.3   Stock standard solutions  must  be replaced after  1  year or
      sooner  if  comparison  with a quality control  reference standard indicates
      a  problem.

      5.3   Intermediate  standard solutions - An intermediate standard solution
should  be  prepared  containing all  the target  analytes  for the  calibration
standard solutions (separate  solutions  for  PAHs and PCBs)  or all the internal
standards for the internal  standard solution.   The recommended concentration is
100 mg/L.

      5.4   GC/MS tuning  standard -  A methylene chloride solution containing 50
mg/L of decafluorotriphenylphosphine (DFTPP)  should be prepared.   Store at -10°C
to -20°C or less when not being used.

      5.5   Matrix  spike  standard -  Prepare a spiking solution in methanol that
contains five or more of the target compounds at 100 mg/L for solid samples.  The
selection of compounds  should  represent the  boiling point  range of the target
compounds.   The stock and intermediate  standards may be prepared as in Section
5.2 and 5.3 or  commercially  prepared certified standards  are also acceptable.
The  standards  must  be  prepared  independently  from  the  calibration  stock
standards.

      5.6   Blank soil used for  the preparation  of the calibration standard soil
and internal  standard soil  is prepared  as outlined below.

           5.6.1    Obtain  a clean (free  of  target  analytes  and  interferences)
      sedimentary soil.  Dry and then grind it in a mortar and pestle.  Sieve the
      ground material through  a 100  mesh sieve.  Several 50 mg aliquots should
      be  extracted  by   TE/GC/MS (or  other   techniques)  to  determine   if  any
      compounds  are present that could interfere with the compounds in Tables 1
      and 2.

           5.6.2   If no interferences  are  found,  300-500 grams of  dried  and
      sieved blank soil  is tumbled for 2  days in a clean glass container with a
      Teflon®-!ined cap to ensure homogeneity before the analytes  are spiked onto
      the soil.

      5.7   Internal standard soil - The internal standard is  prepared on a blank
soil   (Section 5.6).  The  internal  standard soil should contain all  compounds
listed  in  Table  3  at  a  concentration of   50  mg/kg for  each  compound.
Commercially-prepared soil  standards may be  used if they  are certified by  the
manufacturer or by an independent source.

      5.8   Calibration standard  soil  - The calibration standard  is  prepared on
a blank soil  (Section  5.6).    The  calibration  standard soil  must  contain  all
target analytes  to be reported,  at  a concentration of 35 mg/kg for the PAHs  and
10 mg/kg for the PCBs.   The PCBs are prepared at a lower concentration  because
expected concentrations  in soil  are expected  to  be lower. If  preferred, both the
PAHs  and PCBs  may  be  prepared  at the  same concentration.  See Table 1 (PAH)  and
Table  2 (PCB)  for  the  analytes  that  have   been  tested   by  this  method.

                                  8275A - 5                         Revision 1
                                                                  January 1995

-------
Commercially-prepared soil standards may be  used  if  they are certified by the
manufacturer or by an independent source.

      5.9   Preparation  of the  internal  standard and calibration standards on a
blank soil.

           5.9.1    The   50  mg/kg  internal   standard  soil   and both  the  PAH
      calibration  standard  (35 mg/kg) and PCB  Calibration  Standard  (10 mg/kg)
      soils are  all  prepared  by the  same technique.  The intermediate standard
      solutions  (Section 5.3)  or commercially-prepared certified solutions are
      used  for dosing a  weighed  amount of blank soil (Section 5.6).  Weigh 20.0
      g of  blank soil (as prepared in Section  5.6)  into a 4-oz. glass container.
      Water is added  (5% by  weight)  to aid  in  the  mixing and  dispersal  of
      analytes to the more polar sites in the  soil, as  occurs  in nature.  For an
      intermediate  standard containing  100 mg/L of each  compound:   add 10.0 ml
      to the wetted blank soil  for the internal standard soil; add 7.0 ml for the
      PAH  calibration standard soil;  and 2 ml for  the PCB calibration standard
      soil.  Add additional methylene chloride so that the total solvent provides
      a  slight solvent  layer  above the  soil.   This  helps  to  distribute the
      standard compounds homogeneously throughout  the  soil.

           5.9.2    The   solvent  and  water  are  allowed  to  evaporate  at  room
      temperature  until   the soil  appears dry  (usually  overnight).    The  soil
      containers  are tightly  capped with Teflon®-!ined  caps and  placed  on  a
      tumbler that slowly rotates and mixes the  contents.  All soils are tumbled
      for  at least  five  days to  ensure homogeneity.

           5.9.3    Internal standard soil  and calibration standard soil should
      be stored in amber glass vials with Teflon® seal  caps  at -10°C to -20°C or
      less  and protected from  exposure to light and  moisture.  The soil standards
      should  be  stable   for  up  to  90  days  under  these  storage  conditions.
      Internal  standard  and  calibration standard  soils   should  be  checked
      frequently against the calibration  solutions  for signs of degradation.  The
      check is performed by   adding an  equivalent concentration  of  standard
      solution to the frit in  the sample crucible  lid just prior to transfer of
      the  crucible  and lid to  the thermal extraction unit.

           5.9.4    Internal  standard and  calibration standard  soils  must  be
      replaced if the above check indicates degradation.

NOTE:   The more volatile PAHs  and  PCBs in the soil  calibration standards may
        exhibit  higher  concentrations  than  the  calibration solutions.   This
        results  from the possibility of evaporation  losses  from  the crucible
        frit lid of the more  volatile analytes.

      5.10   Methylene chloride,  methanol, carbon disulfide, toluene,  and other
appropriate solvents -  Pesticide quality or  equivalent.


6.0   SAMPLE COLLECTION,  PRESERVATION AND HANDLING

      See the introductory material  to this Chapter, Organic Analytes, Sec. 4.1.
                                   8275A  -  6                         Revision 1
                                                                  January 1995

-------
7.0  PROCEDURE
     7.1  Sample crucible preparation

WARNING:   Do not touch  the  crucibles with your  fingers.   This  can result in a
          serious burn during removal  from the muffle furnace.  Clean crucibles
          can  be contaminated  with oils from the fingers.   Always  handle the
          sample crucibles  and lids with stainless steel tweezers.

          7.1.1   Turn  on the muffle furnace for cleaning crucibles and let it
     heat to 800°C for  at least 30 minutes.

          7.1.2   To  clean  the crucibles,  load the sample  crucibles  and lids
     into the  muffle  furnace tray  and  place  in  the oven.   Leave  in  the muffle
     furnace for 15 minutes then  remove tray  and place on cooling pad (at least
     15-20 minutes) before  transferring crucibles to the "clean" petri dish.

          7.1.3   All sample crucibles and lids should be pre-cleaned and placed
     in a covered petri  dish.  Prepare a sufficient number of crucibles and lids
     to prepare  a  5-point  calibration  curve and/or  for the number  of sample
     analyses  planned.

     7.2  Initial calibration of the TE/GC/MS system

          7.2.1   Set the TE/GC/MS to the following recommended conditions and
     bake out  the system.

     ON-LINE bakeout procedure -This procedure  should be performed before each
set of calibration runs.  If  the autosampler is used, this should be incorporated
into the  autosampling  sequence.

IMPORTANT:    Sample crucible must  be removed from the  thermal  extraction unit
             BEFORE bakeout procedure  begins.   It is  not  necessary  to acquire
             MS data during  a bakeout although GC/MS data should be taken during
             analysis of a method blank (following a bakeout) to monitor system
             contamination.

     GC initial column temp, and hold time:      35°C  for 4  minutes
     GC column temperature program:
     GC final column temperature hold:
     GC cool time:
     GC injection port temperature:

     MS transfer line temperature:
     GC Carrier gas:
     TE transfer line temperature:
     TE interface oven temperature:
     TE helium sweep gas flow rate:
     TE sample chamber heating profile:
35 to 325°C at 20°C/min
325°C for 10 minutes
325°C to 35°C in 4 minutes
335°C;   splitless  mode   for
entire run
290 - 300°C
Helium at 30 cm/sec
310°C
335°C
40 mL/min
Hold 60°C for 2 min;      60 -
650°C in 12  min; hold 650°C for
2 min; cool  to 60°C.
                                  8275A - 7
                    Revision 1
                  January 1995

-------
     7.2.2   Set  the  TE/GC/MS   system  to   the   following   recommended
conditions for calibration and sample analysis assuming  a  30-m  capillary
column (see Section 4.1.5).
Mass range:
MS scan time:
GC initial column temp, and hold
GC column temperature program:
GC final column temperature hold:

GC column cool rate:
GC injector type:
GC
GC
injection port temperature:
injection port setting:
MS transfer line temperature:
MS source temperature:

MS solvent delay time:
MS data acquisition:
Calibration Standard Soil weight:

Carrier gas:
TE transfer line temperature:
TE interface Oven temperature:
TE helium sweep gas flow rate:
TE sample heating profile:
                                                45 - 450 amu
                                                1.0 to 1.4 scan/sec
                                      time:     35°C for 12 minutes
                                                35 - 315°C at 8°C/min
                                                315°C  for  2  min  (or  until
                                                benzo(g,h,i)perylene elutes.
                                                315°C to 35°C in 4 minutes
                                                Split/splitless capillary; 35:1
                                                split ratio
                                                325°C
                                                Splitless  for  30  sec,   then
                                                split mode for remainder of run
                                                290 - 300°C
                                                According   to   manufacturer's
                                                specifications
                                                15 minutes
                                                Off at 49 minutes
                                                See Section 7.2.5.3 for initial
                                                calibration.
                                                Helium at 30 cm/sec
                                                310°C
                                                335°C
                                                40 mL/min
                                                hold  60°C  isothermal  for   2
                                                min; 60 - 340°C in 8 min;  hold
                                                340°C  for  3 min;  cool  340  -
                                                60°C for 4 min.

NOTE:   All calibration standards and samples must  be analyzed  under the  same
        split ratio settings.

           7.2.3    Method blank - A blank should follow the ON-LINE bakeout using
     the conditions listed in Section 7.4.2.  Acquire  the MS data and determine
     that the system is free  of target analytes  and interferences at the project
     required Method  Detection  Limit (MDL).   Make  appropriate  corrections  if
     contamination  is observed  (i.e., bake  out,  change GC column,  change  TE
     sample chamber and/or transfer  line).

           7.2.4    The GC/MS  system must be  hardware tuned to meet  the  DFTPP
     criteria in Method 8270.  Add 350 ng (because of the 35:1 split) of DFTPP
     to the frit in the lid of the crucible  and analyze following the conditions
     in Section 7.2.2.

           7.2.5    Initial  calibration curve -  A  minimum of  five  calibration
     standards should  be run during  the  initial calibration of the system and
     after any maintenance  procedures which may affect system performance.  This
     calibration procedure should also be performed if there is more than a 20%
     drift from the initial calibration curve and  the calibration verification
     unless system maintenance corrects  the problem.  Adjust the injection port
     split  ratio   to   35:1  for  the   following   calibration   standard   soil
                             8275A - 8
                                                           Revision 1
                                                         January 1995

-------
      concentration.   Any future modifications of  the  split ratio require the
      preparation  of  a new initial  calibration  curve  at the new split ratio.

                   7.2.5.1  Using  forceps,  remove  a sample  crucible from the
           clean dish  and place on the  analytical balance.  Tare or establish the
           weight  to  the nearest 0.1 mg and  place on  a  clean surface.

                   7.2.5.2  Weigh 10 mg (±3%) of internal standard soil (Section
           5.7)   into the sample crucible using a fused quartz sample spatula.
           Place  crucible back  on the  balance  and determine  weight.   Record
           current weight and  tare balance for  the  next step.

                   7.2.5.3  Weigh  the  calibration  standard  soil  into  the
           crucible (according to guidance below on  PAHs  and  PCBs) and record
           weight.    Place a  lid  on   the  crucible  and  load  into  the  Thermal
           Extraction  Unit  or  position in  the  autosampler.    All  analysis
           information and conditions  should  be recorded in a  sample  log.

NOTE:   If  commercially-prepared  standards  are used, the  weights may  vary
        slightly  from what are  presented below.  This  is acceptable  as long as
        the calibration  curve is within the  linear range of the  GC/MS system.

           PAH Standard:

           50 mg (±3%) of 35 tug/kg PAH  calibration standard  soil  (Section 5.9).

           Repeat  the  process  with 40, 20,  10,  and 5 mg (±3%) of 35 mg/kg PAH
           calibration standard  soil + 10 mg  of 50  mg/kg IS soil into separate
           crucibles.

           This results  in 50, 40, 20, 10,  and 5 ng  respectively on column of
           each target analyte in  the  calibration standard.

           PCB Standard:

           50 mg (+3%) of 10 mg/kg  PCB  calibration standard soil  (Section 5.9).

           Repeat  the  process  with 40, 20,  10,  and 5 mg (±3%) of 10 mg/kg PCB
           calibration standard  soil + 10 mg  of 50  mg/kg IS soil into separate
           crucibles.

           This results  in 10, 8, 4, 2, and 1 ng respectively on column of each
           target  analyte  in the calibration  standard.

NOTE:   The sensitivity of the GC/MS system may require  adjustment of the above
        standard weights  (calibration   and internal) either up or down.

                   7.2.5.4  A  split ratio of 300  or 400:1  is recommended for
           high concentration  samples.   A new calibration  curve  at the  higher
           split ratio is  required using a calibration standard soil containing
           an  appropriate concentration of  target  analytes  (approximately  10
           times more concentrated  to achieve a similar concentration on column).
                                  8275A - 9                         Revision 1
                                                                  January 1995

-------
     7.2.6   Analysis  - Upon method start, the sample  is  loaded  into  the
fused quartz sample  chamber.   The  sample chamber is heated to 340°C  and
held isothermal  for 3 minutes.  Helium  carrier/sweep gas passes through  the
sample chamber at a flow rate of 40 mL/min. Thermally-extracted compounds
are swept through a  deactivated  fused silica line into the GC injection
port where  they  are split  (low -35:1)  or  (high  -400:1)  before  being
concentrated on  the head of the GC column which is held isothermal  at 35°C.
Once thermal  extraction  is  complete   (13  min.),  the  sample  chamber  is
cooled, the GC oven is heated to 315°C  at a rate of 10°C/min.  (or according
to required separation  needs).  Exact thermal  extraction method parameters
may be adjusted according to separation requirements.

     7.2.7   Calculate response factors (RFs) for each  analyte (using  the
internal standard assignments  given in Tables  4  and 5) and evaluate  the
linearity of the calibration as described in  Sec. 7.0 of Method 8000.

7.3  Calibration verification of the TE/GC/MS system

     7.3.1   Prior to analysis of samples, the  DFTPP tuning  standard must
be analyzed.  Follow the  guidance  in  Section 7.2.4,  The DFTPP  criteria
must be demonstrated during each 12-hour shift.

     7.3.2   At  the  beginning  of each  12-hour shift,  a method  blank  is
analyzed  using   the  conditions  in  Section  7.2.2.    Also,  the  mg   of
calibration standard  soil  used  for the midpoint of the initial  calibration
curve and 10 mg of internal standard  soil are analyzed  and  the RF  values
are calculated  for each  target analyte.  Calculate the % difference  for
each target analyte  as described  in Sec.  7.0 of  Method 8000.  If  the  RF
values of each target analyte  are not  within 20% of  their mean RF  values
determined during the  initial  calibration,  then  the initial  calibration
sequence  must  be  repeated unless  a   calibration  verification  standard
analyzed after system maintenance meets the % difference criteria.

     7.3.3   After every 6 hours of operation,  a  method blank  is  analyzed
to verify that the system is still  clean.

7.4  Sample preparation,  weighing and  loading

     7.4.1   Sample preparation - Decant and  discard any water layer on a
sediment sample.   Discard any  foreign objects such as  pieces   of  wood,
glass,  leaves and rocks.  Sample preparation requires homogenizing the  wet
or dry sample as well  as possible and  selecting  a representative aliquot
for analysis.   Extremely wet  samples  (high  H20  and solvents) can  cause
excessive pressure in the  MS  if too much sample is inserted in  the system.
See Sections  7.4.3.1  and  7.4.3.2  as  guidelines for  sample  weight  and
moisture considerations.

     7.4.2   Determination  of  sample  % dry weight  -  In  certain  cases
involving soil/sediment  samples,  sample results  are  desired  based on  a
dry-weight basis. When such data are desired, a portion of sample  for this
determination should be weighed out at  the same  time as the portion used
for analytical determination.  Also, for any sample that appears to contain
moisture, the %  moisture must be calculated to determine whether drying of


                             8275A  - 10                       Revision 1
                                                            January 1995

-------
      the  sample is  necessary  prior to grinding  in  a mortar  and  pestle (see
      Sections  7.4.3  and  7.4.4).

WARNING:   The  drying oven should be contained in a hood or vented.  Significant
           laboratory contamination  may  result  from  a  heavily  contaminated
           hazardous  waste  sample.,

           Weigh 5-10 g of a portion of sample into a tared crucible.  Determine
      the % dry weight of the sample  by drying overnight at 105°C.  Allow to cool
      in a  desiccator before weighing.   Discard this  portion after weighing as
      a  separate (unheated) portion will always be used for analysis.  Calculate
      the % dry weight as follows:

                      ...     ...    g of dry sample  ,nn
                      % dry weight =	—xlOO
                                       g of sample
                       % moisutre = 100  - (% dry weight)


           7.4.3    Wet Samples  (greater  than  20%  moisture)

                   7.4.3.1   For samples  where naphthalenes are target analytes:

                   Perform   the  following  steps   quickly to  minimize  sample
           exposure to air,  thereby causing possible loss of naphthalenes as well
           as  sample  weight variability  because of loss of moisture.   Tare the
           crucible,  weigh  10  mg  of  IS  soil,  then  add  10-20 mg  of a  wet,
           representative sample  portion.   Record the  sample weight and insert
           the crucible  into the  TE Inlet  system.

                   7.4.3.2   For wet samples  where naphthalenes are not target
           analytes:

                   A  representative aliquot  (3-5 grams) of sample  should  be
           spread  in  a thin  layer  in a clean shallow container and air dried at
           room temperature  (25°C)  in a  hood for  30  -  40 minutes.

                            7.4.3.2.1    Thick layers of clay type sediment may
                   require  longer drying periods.

NOTE:   No heat should be used to  aid drying.

                            7.4.3.2.2    When dry, scrape the sample loose from
                   the container  walls and break  into  uniform particle size or
                   grind  in a mortar and  pestle  until reasonably  uniform and
                   homogeneous  in texture.   Sieve through a  60-mesh  sieve and
                   store  in  a sample vial.
                                  8275A - 11                        Revision 1
                                                                  January 1995

-------
           7.4.4    Dry  samples  (less  than  20 % water)

                   To prepare dry samples,  homogenize 5-10 grams  in a mortar and
           pestle and sieve through a 60-mesh screen  and store in a sample vial.

           7.4.5    Internal  standard  weighing

                   7.4.5.1   Using forceps, remove  a sample crucible  from the
           clean dish and place  on the analytical  balance.   Tare or establish the
           weight to the  nearest 0.1  mg  and place on a clean surface.

                   7.4.5.2   Weigh 10  mg  (±3%) of internal standard soil mixture
           into the sample crucible using a fused quartz sample spatula.  Place
           crucible back  on  the balance and determine weight.   Record current
           weight and tare balance for the next step.

           7.4.6    Sample weighing  - An aliquot  (3  -  250 mg) of  the prepared
     sample  is  removed with a clean fused quartz  spatula  and placed  in the
     sample crucible and its weight  determined.  The weight of the sample to be
     loaded  into  the  thermal   extraction crucible  should  be  determined  as
     follows:

                   7.4.6.1   If  low  levels (0.02  -  5.0  mg/kg  and   low  total
           organic  content)  are expected,  100 to  250 mg  of (dry) sample should
           be weighed (assuming a 35:1 split ratio).

NOTE: As per Section  1.2, the estimated quantitation limit of this method is 1
      mg/kg.  Any concentrations that are determined to  be  lower  than 1 mg/kg
      would be considered estimated  concentrations.

                   7.4.6.2   If  high  levels  (500-1500 mg/kg  and  high  total
           organic  content)  are expected,  3  to  5 mg of  (dry)  sample  should be
           weighed  (assuming a  35:1 split ratio).

                   7.4.6.3   For  intermediate   levels,   adjust   the  weights
           accordingly.

                   7.4.6.4   If  the expected concentration exceeds 1500 mg/kg, a
           greater  split  ratio  is required.   A split ratio  of 300  to  400 is
           recommended.   This,  of course, requires an initial  calibration curve
           developed with the selected split ratio.

                   7.4.6.5   For  samples  of  unknown  concentration  or  total
           organic content, weigh less than 20 mg of  sample for the initial run.

NOTE: It is highly  recommended  that samples of unknown concentration be screened
      prior to  TE/GC/MS  analysis.    This  will  prevent  the  need  to  reanalyze
      samples as well  as  protect  the  system from overload which causes downtime
      while performing  system maintenance.  The screening  may be performed using
      the  optional  FID device  that  is  available as an add-on to  the TE/GC/MS
      device or  by using a  rapid  semiquantitative extraction  with  methylene
      chloride and  injection on a GC/FID to  determine relative concentrations.
                                  8275A - 12                        Revision 1
                                                                  January 1995

-------
                   7.4.6.6   Select a sample for matrix spike determination (see
          Method 3500 for guidance).  Weigh one or two portions into crucibles
          containing  internal  standard  (see  same  Section  cited  above  for
          guidance on whether to analyze a matrix spike duplicate or a duplicate
          sample).   Add 5.0 ^L  of the matrix  spike standard  (Section  5.5)
          directly to the sample,  immediately cover with lid and transfer to the
          thermal  extraction  unit  or the autosampler.

          7.4.7    Loading  sample  -  Make  sample concentration assessment  and
     weigh  sample  into  crucible containing the  previously  weighed  internal
     standard  soil.   Record sample weight (to the nearest  0.1 mg),  cover  the
     crucible  with lid  and  place  covered  crucible  into  the thermal  extraction
     unit or  autosampler tray.   If the sample is wet  and/or  target  compounds
     have a higher volatility than n-dodecane,  the  autosampler tray  should  be
     chilled to 10-15'C.

          7.4.8    Analysis  -  The sample is loaded into the fused  quartz sample
     chamber of the thermal extraction unit.   See Section 7.2.7 for details  on
     the operation of the TE/GC/MS  system.
                   7.4.8.1
          signal  to noise
          appropriate from
For  extremely  low concentration samples where the
ratio  is  less than  3:1,  increase  sample  size as
detector response after repeating Section 7.4.5.
                   7.4.8.2   If too  much  sample  is  extracted and  GC  column
          overloading  is  evident,  bake out  system  (as  in Section  7.2.1)  and
          analyze  a  blank  to  determine  if  additional  system  cleaning  is
          necessary  (Section 7.2.3).   Use  a  smaller aliquot  of  the  sample
          (decreasing  sample  size as required) after repeating Section 7.4.5.

     7.5  Maintenance  bakeout procedure

          7.5.1    System  bakeout  conditions:   For OFF-LINE  (no  autosampling)
     conditions  following  an extremely  overloaded  system   and  for  routine
     cleaning maintenance.

IMPORTANT:  Sample crucible  must  be  removed  from  the  thermal extraction  unit
           BEFORE bakeout  procedure  begins.

          Before this bakeout procedure, the TE interface oven should be cooled
     so that the fused  silica transfer line capillary can be removed.   Following
     the bakeout a new transfer line capillary should be installed.
     GC initial column temp, and hold:
     GC injection port temperature:
     MS transfer line temperature:
     GC Carrier gas:
     TE transfer line temperature:

     TE interface oven temperature:
     TE sweep gas flow rate:
     TE sample chamber heating profile:
                                  8275A  -  13
                    335°C, hold for 20 minutes
                    335°C; set in split mode
                    295 - 305°C
                    Helium at 30 cm/sec
                    OFF;   until   new   capillary
                    installed
                    400°C
                    MAX; approx 60 mL/min;
                    Heat  to  750'C and  hold 700°C
                    for 3 min; cool to 60°C.
                                        Revision 1
                                      January 1995

-------
      7.6   Qualitative  analysis

      Follow  the  procedures  in  Method  8270,   Sec.  7.0,  to  identify  target
compounds.

      7.7   Quantitative analysis

      Identified compounds are quantitated via the  internal  standard calibration
technique  using  the  integrated  abundance from  the   EICP  of  the  primary
characteristic ion.  The internal standard used  should be assigned according to
Table 4.  Calculate the concentration of each identified analyte as follows:

                                    (A )(C. )(W. )
                             f-  _   XX/VIS'VIS'
     where:                        (RF)(A.J(WX)(D)

     Cx  =  Concentration of compound being measured (mg/kg).
     Ax  =  Area of characteristic ion for compound being measured in sample.
     Cis  =  Concentration of internal standard soil (mg/kg).
     WIS  =  Weight of internal  standard soil  (kg).
     Wx  =  Weight of sample (kg).
     RF  =  Mean  response  factor  for  compound  being  measured  from initial
            calibration curve.
     Ais  =  Area of characteristic ion for the internal  standard.
     D   =  (100 - % moisture in sample)/100,  or 1  for wet-weight  basis.
8.0  QUALITY  CONTROL

     8.1   Refer to Chapter One and Method 8000 for specific quality control (QC)
procedures.   Quality control  procedures to  ensure the proper operation of the
various sample preparation and/or  sample introduction techniques can  be  found in
Methods 3500  and 5000.     Each  laboratory  should  maintain a  formal  quality
assurance program.  The laboratory should  also maintain records to document the
quality of the data generated.

     8.2   Quality control  procedures  necessary to  evaluate  the   GC system
operation are found  in  Method  8000, Sec. 7.0 and  include evaluation of retention
time windows,  calibration verification and chromatographic  analysis  of  samples.
Required instrument QC is found in the following sections of Method  8275:

           8.2.1   The   GC/MS  system   must   be  tuned  to  meet   the  DFTPP
     specifications  in Sections 7.2.4  and 7.3.1.

           8.2.2   There  must be  an initial  calibration of the GC/MS system as
     specified in Section 7.2.

           8.2.3   The  GC/MS  system  must  meet  the  calibration  verification
     criteria specified  in  Section  7.3 each  12  hours.
                                  8275A - 14                        Revision 1
                                                                  January 1995

-------
      8.3  Initial   Demonstration  of   Proficiency   -   Each  laboratory  must
demonstrate  initial proficiency with each sample preparation and determinative
method  combination  it  utilizes,  by generating data of acceptable accuracy and
precision for target analytes  in a clean matrix.  The laboratory must  also repeat
the following operations whenever  new staff are trained or significant changes
in instrumentation  are made.  See  Method 8000, Sec. 8.0 for information on how
to accomplish  this  demonstration.   NIST  (National  Institute  of Standards and
Technology) Standard Reference Material  (SRM) #1939 may be used to monitor method
performance  and document data quality.  An SRM with PAHs may be substituted if
PAHs are the primary target analytes.

      8.4  Sample Quality Control for Preparation and Analysis - The laboratory
must also  have  procedures for documenting the effect of  the  matrix on method
performance  (precision,  accuracy,  and  detection limit).   At  a minimum,  this
includes the analysis  of QC  samples  including a  method  blank,  matrix spike,  a
duplicate, and a laboratory control  sample (LCS) in each analytical  batch and the
addition of  surrogates to each field sample and QC sample.

           8.4.1    Documenting the effect of the  matrix  should include  the
      analysis of at least one matrix spike and one duplicate unspiked sample or
     one matrix spike/matrix  spike duplicate  pair.  The decision on whether to
     prepare and  analyze duplicate samples  or a  matrix  spike/matrix  spike
     duplicate must be based on a knowledge of the samples  in the sample batch.
      If samples are expected  to contain target analytes, then laboratories may
     use one matrix spike and a duplicate analysis of an unspiked field sample.
      If samples are not expected to contain target analytes, laboratories should
     use a matrix  spike  and matrix spike  duplicate pair.

           8.4.2    A Laboratory Control  Sample  (LCS) should be included with each
     analytical  batch.   The  LCS consists  of  an  aliquot of  a  clean (control)
     matrix  similar to the sample matrix and of the same weight  or volume.  The
     LCS is  spiked with  the  same  analytes  at the same concentrations  as  the
     matrix  spike.   When the results of the  matrix  spike  analysis  indicate  a
     potential problem due to the sample matrix itself, the LCS results are used
     to verify that the  laboratory can  perform the analysis in a clean matrix.

           8.4.3    See  Method 8000, Sec.  8.0  for the details  on  carrying  out
     sample  quality control procedures  for preparation and analysis.

     8.5  Surrogate recoveries - The laboratory must evaluate surrogate recovery
data from individual samples versus  the surrogate control  limits developed by the
laboratory.  See Method 8000,  Sec.  8.0 for information  on  evaluating surrogate
data and developing and updating surrogate limits.

     8.6   It is  recommended  that the  laboratory adopt  additional  quality
assurance practices for use with this method.   The  specific practices that  are
most productive depend upon the needs of  the  laboratory and  the nature  of  the
samples.  Whenever possible,  the laboratory should  analyze standard reference
materials and participate in relevant performance evaluation studies.
                                  8275A - 15                        Revision 1
                                                                  January 1995

-------
9.0  METHOD  PERFORMANCE

     9.1  Multilaboratory  precision  data for PAHs and for  a  few semivolatile
compounds are presented in Table 5.  The results  are  based  on the analysis of
test soils spiked at  10 mg/kg and analyzed by 3 different laboratories.  A Ruska
ThermEx inlet interfaced to a GC/MS system was utilized to develop the data.  A
total ion chromatogram generated by TE/GC/MS of PAH analysis is shown in Figure
1.

     9.2  Multilaboratory  performance  data  for  PCB  congeners  are presented in
Table 6.   The results are based  on analyses  of NIST Standard Reference Material
(SRM) #1939 using Method 8275A (Reference 1).  A Ruska ThermEx inlet interfaced
to  a GC/MS  system  was utilized  to  develop the  data.   An  ion  chromatogram
generated by TE/GC/MS of PCB congeners  is shown  in Figure  2.


10.0 REFERENCES

1.   Worden,  R.,  "Method 8275A: Quantitative Addendum For SW-846 Method 8275",
     Research  report  to  the  U.S.  Environmental  Protection  Agency;  Ruska
     Laboratories, Inc., Houston, TX,  1993.

2.   Snelling, R., King, D., Belair,  B.,  "Analysis of  PAHs in Soils and Sludges
     Using Thermal Extraction-GC-MS", Application  Note 228-228; Hewlett-Packard
     Co., Wilmington, DE,  1993.

3.   King, D., Belair, B.,  "Analysis  of PCBs in Soils and Sludges Using Thermal
     Extraction-GC-MS",   Application  Note  228-229;  Hewlett-Packard   Co.,
     Wilmington,  DE,  1993.
                                  8275A - 16                        Revision 1
                                                                  January 1995

-------
                              TABLE 1

  PAH/SEMIVOLATILE CALIBRATION STANDARD SOIL  AND QUANTITATION IONS
Compound                           Quantitation Ion
1,2,4-Trichlorobenzene1                  180
Naphthalene                              128
Acenaphthylene                           152
Acenaphthene                             153
Dibenzofuran                             168
Fluorene                                 166
4-Bromophenyl phenyl ether1              248
Hexachlorobenzene1                       284
Phenanthrene                             178
Anthracene                               178
Fluoranthene                             202
Pyrene                                   202
Benzo[a]anthracene                       228
Chrysene                                 228
Benzo[b]fluoranthene                     252
Benzo[k]fluoranthene                     252
Benzo[a]pyrene                           252
Indeno(l,2,3-cd)pyrene                   276
Dibenz[a,h]anthracene                    278
Benzo[g,h,i]perylene                     276
1   This analyte may be deleted if the target analytes are PAHs only.

   All  compounds are present at 35 mg/kg.
                            8275A  -  17                        Revision  1
                                                            January  1995

-------
                                   TABLE  2

                         PCB  CALIBRATION STANDARD SOIL

IUPAC # CAS #
1
11
18
26
31
52
49
44
66
101
118
138
187
128
180
170
194
206
209
2051-60-7
2050-67-1
37680-65-2
3844-81-4
16606-02-3
35693-99-3
41464-40-8
41464-39-5
32598-10-0
37680-73-2
31508-00-6
35065-28-2
52663-68-0
38380-07-3
35065-29-3
35065-30-6
35694-08-7
40186-72-9
2051-24-3
2
3
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
Compound Name Quantitation Ion
-Chlorobiphenyl
,3'
,2'
,3'
,4'
,2'
,2'
,2'
,3'
,2'
,3'
,2'
,2'
,2'
,2'
,2'
,2'
,2'
,2'
-Dichlorobiphenyl
, 5 -Tri chlorobiphenyl
, 5-Tri chl orobi phenyl
,5-Trichlorobiphenyl
,5,5' -Tetrachl orobi phenyl
,4, 5 '-Tetrachl orobi phenyl
,3, 5 '-Tetrachl orobi phenyl
,4,4' -Tetrachl orobi phenyl
, 4, 5, 5' -Pent achl orobi phenyl
,4,4' ,5-Pentachlorobiphenyl
,3,4,4' ,5'-Hexachlorobiphenyl
,3,4' ,5,5' ,6-Heptachlorobiphenyl
,3,3' ,4, 4 '-Hexachl orobi phenyl
,3,4,4' , 5, 5 '-Heptachl orobi phenyl
,3,3' ,4,4' ,5-Heptachlorobiphenyl
,3,3',4,4',5,5'-Octachlorobiphenyl
,3,3' ,4,4' ,5, 5 ',6-Nonachl orobi phenyl
,3,3',4,4',5,5',6,6'-Decachlorobiphenyl
188
222
258
258
258
292
292
292
292
326
326
360
394
360
394
394
430
392
426
All compounds are present at 10.0 mg/kg
                                  8275A - 18
  Revision 1
January 1995

-------
                                    TABLE 3

                            INTERNAL STANDARD SOIL
      Compound                           Quantitation  Ion
2-Fluorobiphenyl                               172
Phenanthrene-din1                               188
              Mo
Benzo[g,h,i]perylene (  C12)                    288
1  This internal standard is more susceptible to  soil microbial degradation.  It
  is suggested that a 13C-labeled phenanthrene be  substituted.
                                    TABLE  4

              INTERNAL STANDARDS WITH CORRESPONDING PAH ANALYTES
                           ASSIGNED FOR  QUANTITATION
2-F1uorobipheny1

  Naphthalene
  Acenaphthylene
  Acenaphthene
  Fluorene
  All PCB Congeners from Table 2
Phenanthrene-dlO

  Phenanthrene
  Anthracene
  Fluoranthene
  Pyrene
Benzo(g,h,i)perylene(  C12)

  Benzo[a]anthracene
  Chrysene
  Benzo[b]fluoranthene
  Benzo[k]fluoranthene
  Benzo[a]pyrene
  Indeno(l,2,3-cd)pyrene
  Oibenz[a,h]anthracene
  Benzo[g,h,i]perylene
                                  8275A - 19                        Revision 1
                                                                  January 1995

-------
TABLE 5
MULTI- LABORATORY PRECISION DATA FROM OF ANALYSIS OF SAMPLES
CONTAINING PAHs/SEMIVOLATILES USING TE/GC/MS3
#
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Compound
l,2,4-Trichlorobenzeneb
Naphthalene
Acenaphthylene
Acenaphthene
Dibenzofuran
Fluorene
4-Bromophenyl phenyl etherb
Hexachlorobenzeneb
Phenanthrene
Anthracene
Fluoranthene
Pyrene
Benzo[a]anthracene
Chrysene
Benzo[b]fluoranthene
Benzo[k]fluoranthene
Benzo[a]pyrene
Indeno(l,2,3-cd)pyrene
Dibenz [a, h] anthracene
Benzo[g,h,i]perylene
Quant. Ion
180
128
152
153
168
166
248
284
178
178
202
202
228
228
252
252
252
276
278
276
Results for 10 mg/kg
Mean
11.0
13.0
10.4
11.1
10.8
11.0
11.1
10.8
10.6
10.9
10.7
10.8
10.1
10.4
10.9
10.4
10.9
11.0
10.5
10.8
RSD (%)
35.7
45.3
18.1
22.6
22.9
24.0
17.6
18.5
20.2
34.4
17.0
19.9
24.1
17.8
23.1
30.0
25.7
21.5
25.8
21.5
a These data are the compiled results of studies done by three different
  laboratories.  These results are from samples spiked at 10 mg/kg.

b Not a PAH.

  Data are taken from Reference 1.
                             8275A  -  20
  Revision 1
January 1995

-------





o
1—
1 •>

QJ
E
QJ
CO

^
QJ
>
'^
*-'
CT>
ro
CTl
1 — 1
CO
2: 2:
CO O
CJ3
CO \
O UJ
O_ | —
1 1 ^_
o CQ
IO CO 2!
I-H ai
LU CO CO
CQ _l CD
< <£ 0
1— Z CL.
-
o£
o
i—
>!
0 l~
Q) S
o: z
o 4*—
s^ o
,_^
CD
_Q NX
CO \
*f 1-*\
^- u
(_> >H-
\
UJ CJ
1— C
o
o
o
C '""^
O CD
° \
h- E1
CO -— '
I-H
z


•
•*~^ cz
r5 O
3 >— '
cr





























-a
3

E
°

O
Q- =«=
i — i

CD
CO
0


O
1 —








i — i
^f
CVI






IO
**•
CO





00
LO
CVI














f—
>^
f—
cu
JC
o.

.a
o
l-
o
r—
JC
o
• 1 —
i-
h-
LO
**
CVI
CVI

CO
I-H


CVI
CTl
o



CO
1 — 1








CVI
VD
LO






0
CVI
TJ-





00
LO
CVI














^_
>^
c
QJ
JC
O-

0
o

o
r—
JC
o
^h






00
•*
Ti-





CVI
CTl
CVI









j—
>^
C
Q)
< —
a.
-r—
f~l
O

0

-C
o
re
s-
^_}
QJ
1—
l
^
LO
LO

CVI
CVI

CVI
LO


CO
CTl
O


CO
CD








CTl
CO
CO






CO
^
CO





CVI
CTl
CVI









p—
>^
c
Q)
f—
CL
• r—
r*i
O
i.
O
f—
JC
o
re
S-
-M
QJ
1—
i
^
LO
^1"

CVI
CVI

CTl


ID
CD


CTl
CO








LO
CT>
CD






r^
CD
i — i





CVI
CTl
CVI









!—~
>^
C
QJ

O-

JD
O
S-
o

JC
o
re
s-
^_j
QJ
1—
l
^
LO
CO

CVI
CVI

5


-
o


o
I-H
1 — 1








CVI
0
1 — 1






CO
CTl
O





CVI
CTl
CVI









P_
>^
C
Q)
JC
Q.
•r—
f->
O
S-
0

JC
o

S-
-M
QJ
1—
i
^
•*
«d-

co
CVI

to
IO


CTl
CD


i — i
i — i
i — i








i — i
CTl
CD






CVI
CO
0





IO
CVI
CO







r~
>^
c~
QJ
JC
Q.
-r—
O
O
s-
0

f~
o
re

c
QJ
CL.
l
^
LO
^.
LO
^r

CVI
CVI

o
r-H


*
O



CO








o
r--
0






^H
LO
o





to
CVI
CO







f—
>^
C
OJ

CL
•i —
J2
O
S-
o
r«-
f
O
re
i ^
c
cu
Q-
l
LO
*.
«.
«*"
•Sf

CO
CVI

CO
i-H
1 — 1


CVI
CD


CVI
O
1 — 1








00
LO
o






I~^
LO
o





0
to
co





,_
>^
c
cu

a.

t~t
o
i-
o

r—
O
re
X
QJ

1
V
LO

*»
*!j"
„
*
co

CVI
CVI

00
CO
i-H


IO
CD
CD


00
r--








^j-
i — i
o






oo
1 — t
CD





^j-
CT1
CO



•^
C
QJ

CL
•r—
r*>
O
1-
0

c—
O

4_>
a.
cu
:r

IO

^
LO
«s
LO
r.
«.
**
CO

CVI
CVI

CO
t-H


CD
0


O
00








00
0
CD






O
I-H
0





o
10
co





F^
>^
C
QJ
-C
CL

O
o
i.
o

c**
O
re
X
QJ

l
•^
*^-
«*
^~
r.
«.
CO
co

CVI
CVI

00
CVI
I-H


LO
o
0


LO
r-








CVI
i-H
O






IO
I — 1
0





^~
cn
CO


_
^
Q)
f—
0.

JD

O
5-
O
, —
JC
a
re
4^
CL
a>
~r~
i
^
LO
<.
LO
«*
•^
«^-
r.
«d-
co

CVI
CVI

0
CO
1 — 1


LO
CD
CD


LO
LO








10
0
o






, — 1
1 — 1
0





^>
CTl
CO


_
C
QJ
e—
CL
•r—
o
i
O
S-
o
f—
JC
Cj
re
4_3
CL
QJ

i
LO

^.
^~
««
^«
^
^
CO
CO

CVI
CVI

0
r-H

 QJ
 X
 O
 co
 QJ
 Q)
CO

 S-
 QJ
 >

Si

 c
 oo
 CQ
 CTl
o:
co
 o

 00
•r—
 00


're

 re

 QJ
 re
        CO
        QJ
        00
        as
        E
        re
        o
        00
        CVI
           -a   •
           C  00
           3  CD
           as
           03
       CT>      0.
       co  JC
       en  +->  c
       i— i  O  O
       =tt=  JO  00
       Qi  O  fO
       CO      Q.

       I—  re  o
       co  -i->  o
       i—i  o
       Z  4->  S-
               o
       14-  QJ l(_
        o  jc
           4-> -O
        oo      cu
       •r-  oo -a
        oo  •!- -a
        re  3  cu

        re 're  cu
            QJ
    Q
    <_)
    UJ
    o
    C3
       O
       s-

           QJ
       00  >
       QJ  r—
       •i-  O
       "O  00
       3  QJ
           S-
           c:
           3
 O ^

"^3 'o
 re  <->
 ._,  QJ
        00

       co
        O r-H
        •i- CO
 1 =
 § ^
 (J ^v

 c o
 re ••-
 QJ 4->
 E U
    re
t— s-
CO -rJ
1-H X
       re ca
       i- O
       -rJ> Q_
       c
       O) T3
       O  C
       c  re
       o
       O CO
          cvi
       c
       re ca
       a> o
       s: CL.
 co

 CQ

 O-

 -o   •
 C r-H
 re
     cu
 CO  O
 CVI  C
     QJ
 ca  s-
 O  QJ
 Q- t*-
     QJ
 (+_ iy
 O

 00  O
 sz  s_
 O 4-

 -(J  C
 re  Q)
 t    ^/
 +->  re
 C. -rJ
 0)
 O  O)
 c  s_
 o  re
 o
     re
 QJ 4->
x:  re
h— Q
                          T-H LO
                             CTl
                           C CTl
                           O .—l

                           00  >,

                           >  rB
                           QJ  3
                          OL  C
            
-------
                                                                                     1
o

-
CQ
                                                                                          00
                                                                                          CO
                                                                                          CO
                                                                                          o
                                                                                          CO
C5

«t
o:
o
o


                                                                                          (O
                                                                                          CM
                                                                                          CM
                                                                                          CM
                                                                                   —*L  oo
                                                                                                                 en
                                                                                                               c en
                                                                                                               O ^H

                                                                                                               W  >,

                                                                                                               >  re
                                                                                                               0)  3
                                                                                                               C£  C
                                                                                                                  fO
                                                                                                               CM
                                                                                                               CM
                                                                                                               If)
                                                                                                               CNJ
                                                                                                               CO
                                                                                                                       i

-------
CVI


LU

Of.
      o
      LjJ
      t/5

      oi
      UJ
      >
      i—I
      a:
      CO
      <_>
      a.
      CTi
      CO
      C/)
      f—
      C/)
      C/)
      O
      o

      ca
      o
      o_
      o

      oo
      >•
      co

      O
       o
       o;
                                                                                                                               .— i  ID
                                                                                                                                  CTl
                                                                                                                                  >-,
                                                                                                                                  i-
                                                                                                                               CD 3
                                                                                                                               oc: c
                                                                                                            IM
                            CM-
                            IO
                                                                                       CM
                                                                                       oo
                                                                                                            00

                                                                                                            N
 a
 ra
 t_
 O)
 o
 «-*
 (0

 E
 o
 k.
J=
O

 C
 o
                                                                                                             (O
                                                                                                             CM
00  CM  ID O
tn  a>  CM u>
CM  CM  n ro

 C   C   C  C
 o   o   o  o

-------
                                      METHOD  8275A
                 SEMIVOLATILE ORGANIC  COMPOUNDS (PAHs  and  PCBs)
                      IN SOILS/SLUDGES AND SOLID  WASTES USING
     THERMAL  EXTRACTION/GAS CHROMATOGRAPHY/MASS  SPECTROMETRY  (TE/GC/MS)
                        7.1 Prepare crucible.
                      7.2.1 Sat the TE/GC/MS
                      system to recommended
                      conditions and bake out
                           the system.
                      7.2.2 Set the TE/GC/MS
                      system to recommended
                      conditions and calibrate.
                     7.2.3 Analyze method blank.
7.2.3 Make
appropriate
corrections.
      .2.
     Is the
  system free
of target analytes
   and inter-
   ferences?
                         7.2.5 Prepare initial
                          calibration curve.
                                           7.3 Verfiy calibration as
                                                appropriate.
                                           7.4.1 Prepare sample.
                                                                   7.4.2 Determine sample
                                                                        % dry weight.
                                                                     7.4.5 Weigh internal
                                                                         standard.
                                                                     7.4.6 Weigh sample.
    7.4.7 - 7.4.8 Load
  sample into the thermal
extraction unit and analyze.
                                           7.6 Calculate the cone.
                                          of the analytes, or perform
                                            a qualitative analysis.
                                                                      (^   Stop   J)
                                       8275A -  24
                                                        Revision 1
                                                      January  1995

-------
                                  METHOD 8515

        COLORIMETRIC SCREENING METHOD FOR TRINITROTOLUENE (TNT)  IN SOIL


1.0  SCOPE AND APPLICATION

      1.1  Method 8515  is  a procedure for screening soil samples to determine
when 2,4,6-trinitrotoluene (TNT, CAS No. 118-96-7)  is present at concentrations
above 1 ppm.

      1.2  Results obtained using this method should be used to locate samples
with TNT concentrations between 1 and 30 ppm.  Extracts of samples reading >30
ppm should be diluted and re-evaluated  in the test.

      1.3   Using the test  kit from which  this  method was developed,  95% of
samples containing 0.7 ppm of TNT or less will produce a negative result.

      1.4  Method 8515 can be used to screen  soil samples for the  presence of TNT
and other chemically related nitroaromatic compounds e.g., dinitrotoluenes (DNT)
and 1,3,5-trinitrobenzene (TNB).  It  should be used as an  indicator or screening
test for the presence of TNT.  Method 8515 does not measure RDX or HMX.

      1.5   This  method  is  restricted to use  by  or under  the  supervision of
trained  analysts.    Each  analyst  must  demonstrate the  ability to  generate
acceptable results with this method.


2.0  SUMMARY OF METHOD

      2.1    Test kits   are  commercially  available  for  this  method.    The
manufacturer's directions should be followed.

      2.2  In general, the method is  performed using an extract of a soil sample.
The sample  is  treated with  color-change reagents  and  is  read in  a  portable
spectrophotometer.   The colorimetric nature of the test is based on the visual
detection of the reaction product that is formed when polynitroaromatic compounds
react with  acetone  by  ketone  substitution in the presence  of  base.   This
substitution product  is  measured at 540 nm  using a spectrophotometer.   The
concentration of TNT in  an unknown  sample  is determined by  evaluating  the
intensity of the color that is developed.


3.0  INTERFERENCES

      3.1  Chemically similar compounds and compounds which  might be expected to
be found  in  conjunction  with  TNT contamination  were tested to  determine  the
concentration required to produce an equivalent TNT  result.  These data are shown
in Table 1.
                                   8515 - 1                         Revision 0
                                                                  January 1995

-------
4.0  APPARATUS AND MATERIALS

      4.1  TNT Soil Test System (EnSys, Inc.),  or equivalent.  Each commercially
available test kit will  supply or specify the  apparatus and materials necessary
for successful completion of the test.

      4.2  UV/Vis Spectrophotometer, Hach DR/2000, or equivalent.


5.0  REAGENTS

      5.1   Each commercially  available  test  kit  will  supply or  specify the
reagents necessary for successful completion of the test.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory  material  to  this chapter,  Organic Analytes,
Section 4.1.

      6.2  See Section 6.0 of Method 8330.

      6.3  Soil samples may be contaminated, and should therefore be considered
hazardous and handled accordingly.


7.0  PROCEDURE

      7.1  Follow the manufacturer's instructions for the test kit being used.
Those test kits used must meet or exceed the  performance specifications indicated
in Tables 2-3.

      7.2  Dry soil samples  in air at room temperature or colder to a constant
weight, being careful  not to expose the samples to direct sunlight.


8.0  QUALITY CONTROL

      8.1   Follow  the manufacturer's  instructions  for  the  quality  control
procedures specific to the test kit being used. Additionally, guidance provided
in Chapter One should be followed.

      8.2   Use  of  replicate analyses,  particularly  when  results  indicate
concentrations  near  the action  level,  is  recommended  to refine  information
gathered with the kit.

      8.3  Do not use test kits past their expiration dates.

      8.4   Use the test kits  within their specified storage  temperature and
operating temperature limits.

      8.5   Verify operation of  the  colorimeter/spectrophotometer by  use of
appropriate standards.


                                   8515 - 2                         Revision 0
                                                                  January  1995

-------
      8.6  Method 8515 is  intended for field or laboratory use.  The appropriate
level of quality  assurance  should  accompany  the  application of this method to
document data quality.


9.0  METHOD PERFORMANCE

      9.1  Inter- and  intra-assay precision data are provided in Tables  2 and 3,
respectively.

      9.2  This method has  been applied  to  a series of soil samples whose TNT
concentration had  been determined by HPLC  (Method 8330).   These  results are
provided in Table 4.   A high degree of correlation was  observed  between  the HPLC
method and the field method.
10.0  REFERENCES

1.    T. F. Jenkins and M. W. Walsh, "Development of Field Screening Methods for
      TNT, 2,4-DNT, and RDX in Soil", Talanta, 1992, 39 (4), 419-428.

2.    T.  F.  Jenkins,  "Development  of  a  Simplified  Field  Method  for  the
      Determination of TNT in Soil",  Special Report 90-38 (November, 1990) USA
      Cold Regions Research and Engineering Laboratory.

3.    TNT Soil Test System Instructions for Use, EnSys, Inc.
11.0  SAFETY

      11.1   Standard precautionary measures  used for handling  other organic
compounds should be sufficient for the  safe handling of the samples, extracts and
standard solutions specified  in this method.  The only  extra caution that should
be taken is when handling the analytical standard neat  material.  Follow Section
7.2 for drying the neat material  at ambient temperature.   If samples are taken
back to the laboratory for analysis by Method 8330, follow the additional safety
procedures specified in that method.
                                   8515 - 3
  Revision 0
January 1995

-------
                        TABLE  1
TNT Soil Test System
Sensitivity to Explosive Compounds
Compound
2,4,6-Trinitrotoluene
2,4-Dinitrotoluene
2,6-Dinitrotoluene
1, 3, 5-Tri nitrobenzene
2-Nitrotoluene
3-Nitrotoluene
4-Nitrotoluene
4-Amino-2,6-
dinitrotoluene
Nitrobenzene
Minimum Sensitivity
(ppm)
1
1.1
0.6
1
>100
>100
>100
>100
>100
                        TABLE  2
Inter-assay Precision of the TNT Soil Test
Spike
Concentration (ppm)
0
5
10
20
Average Result
(ppm ± SD)
0.0 ± 0.2
5.1 ± 0.4
10.1 ± 0.5
20.1 ± 0.8
%RSD
.
7.8%
4 . 5%
4.2%
These data were generated from 22 replicates  in 10 matrices (9 soil
extracts and 1 acetone control).
                       8515 - 4
  Revision 0
January 1995

-------
                        TABLE  3
Intra-assay Precision in the TNT Soil Test
Spike
Concentration (ppm)
0
10
Average Result
(ppm ± SD)
0.0 ± 0.2
10.2 ± 0.2
%RSD
-
1 . 9%
These data were generated from  the  10 ppm TNT control provided with
the EnSys kit.
                       TABLE 4
Comparison of TNT Soil Test System
with Method 8330 (HPLC)
Sample
ID
012
028
022
021
023
024
027
025
026
016
013
015
020
019
Screening
Test Result
(ppm)
18.9
26.2
34
34.6
37.7
56.5
192
120
120
49
174
150
295
712
HPLC
Results
(ppm)
21.5
29.0
25.2
23.8
28.1
58.5
191
110
131
49
175
135
287
719
Does screening test
agree with HPLC
determination?
yes
yes
no
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
                       8515 -  5
  Revision 0
January 1995

-------
                                  METHOD 9078

          SCREENING TEST METHOD FOR POLYCHLORINATED BIPHENYLS IN SOIL
1.0  SCOPE AND APPLICATION

      1.1  The method may be used to determine the amount of PCB
(polychlorinated biphenyl) contamination in soils such as sand, gravel, loam,
sediment, and clay, assuming that PCBs are the sole source of halogens in the
sample.

      1.2  This electrochemical method is designed to provide quantitative
field results over a range of 2 to 2000 jug/g PCBs, significantly cutting down
on the number of samples requiring laboratory testing.

      1.3  Chlorines are removed from the PCB molecule using an organo-sodium
reagent.  The resulting chloride ions are measured using a chloride specific
electrode.  Analysts must identify the type of Aroclor contamination in order
to use this as a quantitative method.

      1.4   This method is restricted to use by or under the supervision of
trained analysts.  Each analyst must demonstrate the ability to generate
acceptable results with this method.


2.0  SUMMARY OF METHOD

      2.1  A sample of the soil to be tested is extracted with a hydrocarbon
based solvent.  The resulting extract is filtered to remove moisture and
inorganic salts.  The dried extract is reacted with metallic sodium and a
catalyst to strip chloride from any PCB that may be present.  The resulting
chloride ions are extracted into an aqueous buffer solution where they are
detected using a chloride ion specific electrode.

CAUTION:    Some of the reagents used with this testing procedure contain
            flammable solvents, dilute acids, and metallic sodium.   Wear
            gloves and safety glasses while performing tests.  Read all MSDS
            and warnings included with the instrument before starting testing
            procedure.


3.0  INTERFERENCES

      3.1  This procedure is sensitive to any chlorinated compound that is
preferentially soluble in a non-polar solvent.   When analyzing for PCBs, the
presence of other chlorinated organics will result in a high bias.   Iodine and
bromine containing compounds will affect results if present in significant
quantities.   Wet or dry samples may be run, but results for all samples are
calculated on a wet-weight basis.  In one evaluation study (Table 1), 10% of
the measurements were false negatives.
                                   9078 - 1                         Revision 0
                                                                  January 1995

-------
      3.2  Inorganic chlorides should not interfere using this method if the
sample is extracted with organic solvent.


4.0  APPARATUS AND MATERIALS

      4.1  Electrochemical PCB test kit: L2000® PCB/Chloride Analyzer, (Dexsil
Corporation, One Hamden Park Drive, Hamden, CT), or equivalent.  Each
commercially available test kit will supply or specify the apparatus and
materials necessary for successful completion of the test.


5.0  REAGENTS

      5.1  Each commercially available test kit will supply or specify the
reagents necessary for successful completion of the test.  Reagents should be
labeled with appropriate expiration dates.


6.0  SAMPLE COLLECTION AND HANDLING

      6.1  See the introductory material to this chapter, Organic Analytes,
Section 4.1.

      6.2  Soil samples may be contaminated, and should therefore be
considered hazardous and handled accordingly.  All samples should be collected
using a sampling plan that addresses the considerations discussed in Chapter
Nine.

      6.3  To achieve accurate analyses, soil samples should be well
homogenized prior to testing.  PCBs are generally not evenly distributed in a
soil sample and extensive mixing must be done to assure consistency.


7.0  PROCEDURE

      7.1  Follow the manufacturer's instructions for the test kit being used.
Those test kits used must meet or exceed the performance specifications
indicated in Tables 1 and 2.


8.0  QUALITY CONTROL

      8.1  Follow the manufacturer's instructions for quality control
procedures specific to the test kit used.  Additionally, guidance provided in
Chapter One should be followed.

      8.2  Use of replicate analyses, particularly when results indicate
concentrations near the action level, is recommended to refine information
gathered with the kit.

      8.3  Method 9078 is intended for field or laboratory use.  The
appropriate level of quality assurance should accompany the application of
this method to document data quality.

                                   9078 - 2                         Revision 0
                                                                  January 1995

-------
9.0  METHOD PERFORMANCE

      9.1  146 soil samples from a PCB contaminated site were analyzed.  There
were 114 individual samples and 32 field duplicates.  Each sample was analyzed
using both the L2000 and GC/MS.  The L2000 analyses were performed on-site in
a mobile lab and the PCBs were analyzed as Aroclor 1242.  Laboratory analyses
were performed on splits of the same samples.  The results from the analyses
are presented in Table 1.

      9.2  After applying accepted statistical methods to account for the
detection limit difference between the two methods the data were evaluated to
determine the acceptability of the L2000 method. A matched-pair students t-
test performed on the L2000 and CLP GC/MS data results in a t value of 0.2141.
This is well below the critical value (1.645 @ 0.05) for rejecting the null
hypothesis indicating that there is no statistical difference between the data
pairs.   An analysis of the data for outliers identified only 2 data points
whose residuals were greater than 3 standard deviations (10 and 5
respectively).  Both points were determined to be in error using other
evidence and were eliminated from the data set.  A linear regression analysis
of the remaining data results in a correlation coefficient of 0.95 and a
positive intercept of 10.98 M9/9-  The slope of 0.985 was not statistically
different from 1 and the intercept was not statistically different from 0.

      9.3  The relative percent difference (RPD) calculated from all valid
duplicates greater than the L2000 detection limit of 2 /xg/g for each method
resulted in a mean RPD of 19% for the L2000 data and a mean RPD of 43% for the
CLP GC-MS method.  A Dunnett's test shows that this is statistically
significant.

      9.4  In a second study, soil samples contaminated with Aroclor 1260 were
taken during a site cleanup.  The samples were split and sent for lab analysis
by Method 8082 as well as analysis by the L2000 in the field.  The results are
reported in Table 2.  A linear regression analysis of the data resulted in a
correlation coefficient of 0.995, a slope of 1.048 and an intercept of -1.48
/Ltg/g indicating that the L2000 is accurate compared to the lab method.  A
calculation of the relative percent difference for data, where duplicates were
run within a method, results in a lower RPD for the L2000 indicating a tighter
data spread and better repeatability.


10.0  REFERENCE

1.    Griffin, Roger D.  Application of a New PCB Field Analysis Technique for
      Site Assessment.  Proceedings of Hazmacon '92 March - April  1992.
                                   9078 -  3                         Revision  0
                                                                  January 1995

-------
                 TABLE  1
L2000 AND LABORATORY METHODS FROM STUDY #1
            Summary of Results
Sample
Number
1
3
4
6
7
8
9
10
11
15
15D
16
17
18
19
23
25
32
33
34
36
38
40
43
43D
50
50D
52
53
54
55
59
60
60D
61
62
L2000
(M9/9)
ND
ND
23.6
ND
ND
3.9
6.9
5.1
2.7
9.4
12.5
484
6.5
382
71.1
48.8
3.5
36
ND
14.4
>2000
778
5.7
4.1
3.6
ND
ND
9.3
25.7
5.1
4.4
ND
2.3
4.4
549
111
GC/MS
(M9/9)
593
114
6.71
67
552
2
1.3
0.172
1.15
9.13
9.84
2110
2.55
45.4
6.7
20.8
11.7
47.6
6
34
816
1030
4.25
1.69
1.74
3.6
4.4
4.21
0.958
0.516
2.4
7.9
0.624
0.577
580
2.35
Results
Agree?
False Neg.
False Neg.
Yes
False Neg.
False Neg.
Yes
Yes
False Pos.
Yes
Yes
Yes
Yes
Yes
Yes
False Pos.
Yes
Yes
Yes
False Neg.
Yes
Yes
Yes
Yes
Yes
Yes
False Neg.
False Neg.
Yes
False Pos.
False Pos.
Yes
False Neg.
Yes
Yes
Yes
False Pos.
                 9078 - 4
  Revision 0
January 1995

-------
Sample
Number
64
65
66
67
68
69
69D
73
74
75
76
78
79
80
84
84D
85
85D
88
88D
89
90
90D
91
91D
92
92D
95
95D
100
100D
101
102
102D
103
104
107
108
109
109D
L2000
Ug/g) _
172
ND
2.1
7.5
8
5.8
4.4
37
22
61
82
21
148
ND
7.6
10.9
593
596
ND
ND
ND
2
ND
1650
1608
3.14
3.4
20.6
20.1
384
363
8.3
6.3
5
75.2
4.1
161
6.1
P
10.3
GC/MS
(M9/9)
19
3.1
1.98
0.081
0.504
ND
ND
15.8
13.3
23
46.7
2.27
42.8
3.8
1.16
1.08
428
465
2.7
1.77
45
1.01
1.4
1630
1704
1.21
ND
17.5
31.2
177
167
1.21
293
1.77
40.3
7.66
14.1
3.84
ND
ND
Results
Agree?
Yes
False Neg.
Yes
False Pos.
False Pos.
False Pos.
False Pos.
Yes
Yes
Yes
Yes
Yes
Yes
False Neg.
Yes
False Pos.
Yes
Yes
False Neg.
Yes
False Neg.
Yes
Yes
Yes
Yes
Yes
False Pos.
yes
Yes
Yes
Yes
Yes
False Neg.
Yes
Yes
Yes
False Pos.
Yes
False Pos.
False Pos.
9078 - 5
  Revision 0
January 1995

-------
Sample
Number
111
112
113
114
L2000
(M9/9)
20
240
21.8
107
GC/MS
(M9/9)
ND
315
14.9
66.3
Results
Agree?
False Pos.
Yes
Yes
Yes
      79 out of 114 samples are reported in Table 1.  Samples that were found
      to be ND for both the L2000 kit and the GC/MS determination were not
      reported.  False negatives and positives were determined based on NDs
      found by each technique relative to the detection limit of the L2000 kit
      (2 M9/g)-  False negatives and positives were also identified when
      concentrations differed by more than an order of magnitude for results
      from the two techniques.
ND = Not detected

15 False positives:
13 False negatives:
67 Non-detects:
  ND - 14.1 ppm by GC/MS
2.7 - 593 ppm by GC/MS
  ND - 2.5 ppm by GC/MS
                                    TABLE  2

                     L2000 AND METHODS 8082 FROM STUDY #2
                              Summary of Results
Sample
Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
Method 8082
(M9/9)
83
21
12
300/375
29
106/134
3
9.3
1.5
99
7/9
3.6
4.2/6.2
290
L2000 Results
L_ (M9/9)
79/76
22
14
357/326/327
27
116/117
7.6
7.2
5.2
93
13
12
2.9
L_ 254/265
                                   9078 - 6
                                              Revision 0
                                            January 1995

-------
                                  METHOD 9079

              SCREENING TEST METHOD FOR POLYCHLORINATED  BIPHENYLS
                              IN TRANSFORMER OIL
1.0  SCOPE AND APPLICATION

      1.1  Method 9079 may be used to screen hydrocarbon based electrical
insulating fluids for polychlorinated biphenyls (PCBs) at preset levels of 20,
50, 100, or 500 M9/9-  The method is designed to provide screening data
outside of a laboratory environment in under 10 minutes, providing a
colorimetric indication that the concentration of PCBs is above or below the
fixed end point.  Screening procedures may significantly reduce the number of
samples requiring laboratory testing.

      1.2  Chlorines are removed from the PCB molecule using an organo-sodium
reagent.  The resulting chloride ions are measured using a chloride specific
electrode.  Analysts must identify the type of Aroclor contamination in order
to use this as a quantitative method.

      1.3   This method is restricted to use by or under the supervision of
trained analysts.  Each analyst must demonstrate the ability to generate
acceptable results with this method.


2.0  SUMMARY OF METHOD

      2.1  A sample of the oil to be tested is reacted with a mixture of
metallic sodium catalyzed with naphthalene and diglyme at ambient temperature.
This process converts all organic halogens to their respective sodium halides.
All halides in the treated mixture,  including those present prior to the
reaction, are then extracted into an aqueous buffer, a premeasured amount of
mercuric nitrate is added, followed by a solution of diphenylcarbazone as the
indicator.

      2.2  The color of the solution at the end of the test indicates whether
the sample is above or below the preset chlorine level.  A yellow end point
indicates a concentration greater than the set point of the test and a blue-
violet end point indicates a concentration less than the set point of the
test.

      2.3  The end point at which each of the test kits turns positive is
calibrated using Aroclor 1242 standards.  Aroclor 1242 provides a conservative
end point due to its low chlorine content relative to the other Aroclors used
in electrical equipment.  A list of Aroclors used in electrical equipment and
the PCB concentration that gives a positive indication using the 50 /zg/g test
kit is given in Table 1.
                                   9079 - 1                         Revision 0
                                                                  January 1995

-------
3.0  INTERFERENCES

      3.1  Water present in the sample at more than 2% may cause a low
reading.  Water present at this high a "level results in an obvious change in
the sodium reaction and the user should stop the test.

      3.2  High sulfur levels (4%) will cause a high bias possibly resulting
in a false positive reading.  The sample will also smell strongly of sulfur
after the sodium reaction.

      3.3  Any chlorine contained in the sample will be measured as PCB
possibly resulting in a false positive if the total non-PCB chlorine
concentration in the sample is greater than the preset end point for the kit.


4.0  APPARATUS AND MATERIALS

      4.1  Colorimetric test kit: Clor-N-Oil® (Dexsil  Corporation, One Hamden
Park Drive, Hamden, CT), or equivalent.  Each commercially available test kit
will supply or specify the apparatus and materials necessary for successful
completion of the test.  Reagents should be labeled with appropriate
expiration dates.


5.0  REAGENTS

      5.1  Each commercially available test kit will supply or specify the
reagents necessary for successful completion of the test.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1  See the introductory material to this chapter, Organic Analytes,
Section 4.1.

      6.2  Soil and oil samples may be contaminated, and should therefore be
considered hazardous and handled accordingly.  All samples should be collected
using a sampling plan that addresses the considerations discussed in Chapter
Nine.
7.0  PROCEDURE

      7.1  Follow the manufacturer's instructions for the test kit being used.
Those test kits used must meet or exceed the performance specifications
indicated in Tables 1-4.

CAUTION:    Some of the reagents used with this testing procedure contain
            flammable solvents, dilute acids, and metallic sodium.  Wear
            gloves and safety glasses while performing test.  Read all MSDS
            and warnings included with the kit before starting testing
            procedure.

WARNING:    Mercury waste must be properly disposed.

                                   9079 - 2                         Revision 0
                                                                  January 1995
i

-------
8.0  QUALITY CONTROL

      8.1  Follow the manufacturer's instructions for quality control
procedures specific to the test kit used.  Additionally, guidance provided in
Chapter One should be followed.

      8.2  Use of replicate analyses, particularly when results indicate
concentrations near the action level, is recommended to refine information
gathered with the kit.

      8.6  Method 9079 is intended for field or laboratory use.  The
appropriate level of quality assurance should accompany the application of
this method to document data quality.


9.0  METHOD PERFORMANCE

      9.1  A double blind study was conducted using the Clor-N-Oil  50 on 6
spiked transformer oil samples.  The spiking concentrations were chosen to be
identical to those from three sets of the EPA's Water Pollution (WP)
Performance Evaluation Program.  They were made in Shell Diala A electrical
insulating fluid at the concentrations listed in Table 1.

            9.1.1  A total of 38 operators participated in the test, 10 of
      whom had prior experience with the test method, and 27 of whom had no
      prior experience using the kits, and 1 of whom had viewed a test
      demonstration before running the test themselves.  This distribution of
      operators was chosen to verify the robust nature of the method in light
      of the vast range of experience typical of users in the field.

            9.1.2  Each operator was given six random samples containing an
      unknown concentration of PCBs.  Operators recorded their results as
      greater than or less than 50 M9/9-  Only after all of the tests were
      run were the data collected and compared with the known values.

            9.1.3  Out of the 228 tests run, 4 were invalid due to  spillage  of
      reagents or improper kit operation resulting in an incomplete test.  The
      test data are presented in Table 2.  From these data, it is evident that
      there is a much higher likelihood of obtaining a false positive reading
      than a false negative.

            9.1.4  The expected certainties estimated from these data are
      presented in Table 3.  The likelihood of obtaining a false positive
      approaches 90% at 90% of the action level.   This reflects the
      conservative design of the test.  At the action level of 50 M9/9>
      nearly 99% of the samples would be identified correctly as containing  50
      /ug/g of PCB or greater.  These results represent errors due to all
      sources and therefore represent real world performance of the method by
      field personnel.
            9.1.5
      ppm.
ie i .

 For  samples containing 45 /ig/g, 37  of  41  gave  results  >50
                                   9079 -  3                         Revision  0
                                                                  January 1995

-------
      9.2  The real world performance was investigated in detail by Utah Power
& Light Company.  They tested the insulating fluid from approximately 200,000
pieces of electrical equipment with the Clor-N-Oil 50 test kit in order to
classify them as either PCB or non-PCB (i.e.,  containing more or less than
50 jzg/g)-  Each piece of equipment was tested once with the test kit.  A
random sample of 937 of the Clor-N-Oil negatives were tested by Gas
Chromatography to confirm the results.  The results from the testing are
summarized in Table 4.  The false negative rate predicted from the data is
less than 1%  (0.65%).  This means that a transformer can be classified as
non-PCB with greater than 99% confidence.
10.0  REFERENCES

1.    Finch, S.; Lynn, T.B; Lynn T.D.; and Scott, R.P.W., Which Method is the
      Most Reliable in the Field, EPRI PCB seminar Proceedings October 1991.

2.    Mills, David W.; and Rhoads, Kirt W., Clor-N-Oil Test Kit as a PCB
      Screening Tool  1985. EPRI PCB Seminar Proceedings March 1986.  EPRI
      CS/EA/EL-4480 Product 2028.

3.    Rhoads, Kirt W., Clor-N-Oil Test Kit as a Risk Management Tool - An
      Update 1987.  EPRI PCB Seminar Proceedings December 1987.
                                                                                    f
                                   9079 - 4
  Revision 0
January 1995

-------
 TABLE 1
SPIKING CONCENTRATIONS FOR Clor-N-Oil
PCB Concentrations in Shell Diala A
Aroclor
1242
1242
1254
1254
1260
1260
Concentration
21.2 M9/9
45.0 Mg/g
26.3 /xg/g
50.0 //g/g
8.2 M9/9
50.0 M9/9
 TABLE  2
ESTIMATED ERROR RATES FOR Clor-N-Oil 50
PCB Concentration
M9/9
8
21
26
45
50
Expected False
Positive Rate %
<3
2.8
35
88
—
Expected False
Negative Rate %
—
—
—
—
1.4
 TABLE  3
GC RETEST RESULTS OF 930 NEGATIVES CLASS INTERVAL (ng/g)
Number
Percent
0-1
768
82.6
1-5
93
10
6-15
48
5.2
16-25
5
0.5
26-46
10
1.1
47-99
3
0.3
100-475
3
0.3
476-999
0
0
1000+
0
0
9079 - 5
  Revision 0
January 1995

-------
              TABLE 4

TEST RESULTS FOR Clor-N-Oil  50 STUDY
Sample
ID
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
Concentration
PCB (Mg/g)
50
26.3
50
45
21
8.2
21
45
26.3
50
8.2
50
26.3
45
8.2
45
50
50
8.2
26.3
45
21
45
50
21
50
50
Aroclor
1254
1254
1260
1242
1242
1260
1242
1242
1254
1254
1260
1260
1254
1242
1260
1242
1260
1254
1260
1254
1242
1242
1242
1260
1242
1260
1254
Field Test
Result (jug/g)
>50
<50
NA
>50
<50
<50
<50
<50
<50
>50
<50
>50
<50
>50
<50
>50
>50
>50
<50
<50
>50
<50
<50
>50
<50
>50
>50
              9079 - 6
  Revision 0
January 1995
i

-------
Sample
ID
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
Concentration
PCB (/xg/g)
21
45
50
8.2
50
50
50
45
26.3
50
26.3
50
21
50
8.2
26.3
8.2
21
26.3
8.2
21
50
50
8.2
45
21
26.3
21
45
8.2
Aroclor
1242
1242
1254
1260
1254
1260
1254
1242
1254
1254
1254
1260
1242
1260
1260
1254
1260
1242
1254
1260
1242
1260
1260
1260
1242
1242
1254
1242
1242
1260
Field Test
Result (jug/g)
<50
>50
>50
<50
>50
>50
>50
>50
<50
>50
<50
>50
<50
>50
<50
<50
<50
<50
<50
<50
<50
>50
>50
<50
<50
<50
<50
<50
>50
<50
9079 - 7
  Revision 0
January 1995

-------
Sample
ID
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
Concentration
PCB (Mg/g)
50
26.3
21
8.2
45
50
50
50
50
21
50
50
45
26.3
21
8.2
8.2
50
45
45
21
8.2
50
50
26.3
26.3
26.3
21
26.3
45
Aroclor
1254
1254
1242
1260
1242
1254
1254
1254
1260
1242
1260
1260
1242
1254
1242
1260
1260
1260
1242
1242
1242
1260
1254
1254
1254
1254
1254
1242
1254
1242
Field Test
Result (/Ltg/g)
>50
>50
<50
<50
>50
>50
>50
>50
>50
<50
>50
>50
<50
<50
<50
<50
<50
>50
<50
>50
<50
<50
>50
>50
<50
<50
<50
<50
<50
>50
9079 - 8
  Revision 0
January 1995

-------
Sample
ID
88
89
90
91
92
93
97
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
Concentration
PCB (Mg/g)
50
8.2
50
45
8.2
50
45
21
50
45
45
45
26.3
50
26.3
26.3
21
26.3
50
8.2
21
26.3
50
21
45
26.3
50
21
21
26.3
Aroclor
1254
1260
1260
1242
1260
1260
1242
1242
1254
1242
1242
1242
1254
1260
1254
1254
1242
1254
1260
1260
1242
1254
1254
1242
1242
1254
1260
1242
1242
1254
Field Test
Result (MQ/Q)
>50
<50
>50
>50
<50
>50
>50
<50
>50
>50
>50
>50
>50
>50
>50
<50
<50
>50
>50
<50
<50
<50
>50
<50
>50
<50
>50
<50
<50
<50
9079 - 9
  Revision 0
January 1995

-------
Sample
ID
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
Concentration
PCB (Mg/g)
8.2
50
50
50
26.3
45
50
8.2
50
8.2
45
50
21
26.3
50
26.3
45
26.3
50
50
45
21
50
8.2
8.2
50
8.2
45
8.2
26.3
Aroclor
1260
1254
1254
1254
1254
1242
1260
1260
1254
1260
1242
1260
1242
1254
1260
1254
1242
1254
1260
1260
1242
1242
1254
1260
1260
1260
1260
1242
1260
1254
Field Test
Result (jug/g)
<50
>50
>50
>50
>50
>50
>50
<50
>50
<50
>50
>50
<50
<50
>50
<50
>50
<50
>50
>50
>50
<50
>50
<50
<50
>50
<50
>50
<50
<50
9079 - 10
  Revision 0
January 1995
i

-------
Sample
ID
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
Concentration
PCB (/ig/g)
21
8.2
21
50
8.2
50
50
21
50
26.3
21
26.3
8.2
50
45
50
50
50
21
50
50
45
50
50
21
50
8.2
45
8.2
8.2
Aroclor
1242
1260
1242
1254
1260
1254
1254
1242
1260
1254
1242
1254
1260
1254
1242
1254
1254
1260
1242
1254
1260
1242
1260
1260
1242
1260
1260
1242
1260
1260
Field Test
Result (/Ltg/g)
<50
<50
<50
>50
<50
>50
>50
NA
NA
<50
<50
>50
<50
>50
>50
>50
>50
>50
<50
>50
>50
>50
<50
>50
<50
>50
<50
>50
<50
<50
9079 - 11
  Revision 0
January 1995

-------
Sample
ID
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
Concentration
PCB (Mg/g)
45
26.3
50
45
45
21
26.3
50
21
50
50
26.3
8.2
21
8.2
45
8.2
21
26.3
50
50
26.3
45
21
8.2
21
45
50
50
26.3
Aroclor
1242
1254
1254
1242
1242
1242
1254
1254
1242
126
1254
1254
1260
1242
1260
1242
1260
1242
1254
1260
1260
1254
1242
1242
1260
1242
1242
1254
1254
1254
Field Test
Result (jug/g}
>50
>50
>50
>50
>50
<50
>50
>50
<50
>50
>50
<50
<50
<50
<50
>50
<50
<50
>50
>50
>50
>50
>50
>50
<50
<50
>50
>50
>50
NA
9079 - 12
  Revision 0
January 1995
i

-------
Sample
ID
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
Concentration
PCB (Mg/g)
50
50
50
45
21
50
50
8.2
8.2
50
45
26.3
26.3
21
45
45
45
26.3
8.2
45
50
Aroclor
1254
1254
1260
1242
1242
1260
1260
1260
1260
1260
1242
1254
1254
1242
1242
1242
1242
1254
1260
1242
1254
Field Test
Result Ug/g)
>50
>50
>50
>50
<50
>50
>50
<50
<50
>50
>50
>50
>50
>50
>50
>50
>50
>50
<50
>50
>50
9079 - 13
  Revision 0
January 1995

-------
                           CHAPTER FIVE

                    MISCELLANEOUS TEST METHODS
The following methods are found in Chapter Five:
      Method 5050:
      Method 9010A:
      Method 9012A:

      Method 9013:
      Method 9020B:
      Method 9021:
      Method 9022:

      Method 9023:
      Method 9030A:
      Method 9031:
      Method 9035:
      Method 9036:

      Method 9038:
      Method 9056:

      Method 9057:
      Method 9060:
      Method 9065:

      Method 9066:

      Method 9067:

      Method 9070:

      Method 9071A:

      Method 9075:


      Method 9076:
Bomb Preparation Method for Solid Waste
Total and Amenable Cyanide (Colorimetric, Manual)
Total   and  Amenable   Cyanide   (Colorimetric,
Automated UV)
Cyanide Extraction Procedure for Solids and Oils
Total Organic Hal ides (TOX)
Purgeable Organic Hal ides (POX)
Total Organic Hal ides (TOX) by Neutron Activation
Analysis
Extractable Organic Hal ides (EOX) in Solids
Acid-Soluble and Acid-Insoluble Sulfides
Extractable Sulfides
Sulfate (Colorimetric, Automated, Chloranilate)
Sulfate  (Colorimetric,  Automated,  Methylthymol
Blue, AA II)
Sulfate (Turbidimetric)
Determination   of    Inorganic   Anions
Chromatography
Determination of Chloride from HC1/HCL
Sampling Train  (Methods 0050  and 0051)
Chromatography
Total Organic Carbon
Phenolics (Spectrophotometric, Manual 4-AAP
Distillation)
Phenolics  (Colorimetric,  Automated  4-AAP  with
Distillation)
Phenolics    (Spectrophotometric,
Distillation)
Total  Recoverable   Oil  &  Grease
Separatory Funnel Extraction)
Oil and Grease  Extraction Method
Sediment Samples
                       Chlorine  in New and Used
                       by   X-Ray   Fluorescence
                                         by   Ion

                                        Emission
                                        by An ion
                                            with
                                    MBTH    with

                                   (Gravimetric,

                                  for  Sludge and
Test Method  for  Total
Petroleum   Products
Spectrometry (XRF)
Test Method  for  Total
Petroleum  Products  by
Microcoulometry
                       Chlorine in  New and Used
                       Oxidative  Combustion and
                             FIVE - 1
                                    Revision 2
                                    January 1995

-------
Method 9077:

      Method A:
      Method B:

      Method C:

Method 9131:

Method 9132:
Method 9210:

Method 9211:


Method 9212:

Method 9213:


Method 9214:

Method 9215:
Method 9250:
Method 9251:

Method 9253:
Method 9320:
Test Methods for Total  Chlorine  in  New and Used
Petroleum Products (Field Test Kit Methods)
      Fixed End Point Test Kit Method
      Reverse  Titration Quantitative  End  Point
      Test Kit Method
Direct Titration Quantitative End Point Test Kit
Method
Total  Coliform:    Multiple  Tube  Fermentation
Technique
Total Coliform:  Membrane Filter Technique
Potentiometric   Determination   of   Nitrate   in
Aqueous Samples with Ion-Selective Electrode
Potentiometric   Determination   of   Solubilized
Bromide  in Aqueous  Samples with  Ion-Selective
Electrode
Potentiometric  Determination  of
Aqueous Samples with Ion-Selective
Potentiometric   Determination
Cyanide  in Aqueous  Samples  and
Ion-Selective Electrode
Potentiometric  Determination  of   Fluoride  in
Aqueous Samples with Ion-Selective Electrode
Potentiometric   Determination   of   Sulfide   in
Aqueous   Samples   and  Distillates   with   Ion-
Selective Electrode
Chloride (Colorimetric, Automated Ferricyanide AAI)
Chloride  (Colorimetric, Automated  Ferricyanide
AAI I)
Chloride  (Titrimetric,  Silver Nitrate)
Radium-228
    Chloride  in
   Electrode
 of   Solubilized
Distillates with
                        FIVE  - 2
                                    Revision  2
                                    January  1995

-------
                                 METHOD 9012A

            TOTAL AND AMENABLE CYANIDE  (COLORIMETRIC,  AUTOMATED UV)
1.0   SCOPE AND APPLICATION

      1.1    Method  9012  is used to determine the  concentration  of inorganic
cyanide (CAS Registry Number 57-12-5) in wastes or leachate.  The method detects
inorganic cyanides that are present as either soluble salts or  complexes.  It is
used  to determine  values  for both  total  cyanide  and  cyanide  amenable  to
chlorination.  The "reactive" cyanide content of a waste, that is, the cyanide
content that could generate toxic fumes when exposed to mild acidic conditions,
is not distilled by Method 9012 (refer to Chapter Seven).  However, Method 9012
may be used to quantify the concentration of cyanide from the reactivity test.

2.0   SUMMARY OF METHOD

      2.1    The  cyanide,  as  hydrocyanic acid  (HCN),  is  released  from samples
containing  cyanide  by means  of  a reflux-distillation operation  under acidic
conditions and absorbed in a scrubber containing sodium hydroxide solution.  The
cyanide  ion in  the   absorbing  solution  is  then determined  by  automated  UV
colorimetry.

      2.2    In the automated colorimetric measurement, the cyanide is converted
to cyanogen chloride  (CNC1)  by reaction with Chloramine-T at  a pH less than 8
without hydrolyzing to the  cyanate.  After the  reaction  is complete,  color is
formed on the addition of pyridine-barbituric acid reagent.  The concentration
of NaOH  must be  the  same in the standards,  the scrubber  solutions,  and  any
dilution of the  original  scrubber  solution  to obtain  colors of  comparable
intensity.

3.0   INTERFERENCES

      3.1    Interferences  are eliminated or reduced by using the  distillation
procedure.   Chlorine and sulfide are interferences in Method 9012.

      3.2    Oxidizing  agents  such  as  chlorine  decompose  most  cyanides.
Chlorine interferences can be removed by adding an excess of sodium arsenite to
the waste prior to preservation and storage  of the sample  to reduce the chlorine
to chloride which does not interfere.

      3.3    Sulfide interference can be removed  by adding an excess of bismuth
nitrate to  the  waste  (to precipitate  the sulfide) before distillation.   Samples
that contain  hydrogen sulfide,  metal  sulfides,  or  other compounds that  may
produce  hydrogen  sulfide  during  the distillation  should  be  treated by  the
addition of bismuth nitrate.

      3.4    High  results may be  obtained  for  samples  that  contain  nitrate
and/or nitrite.  During the distillation, nitrate and nitrite will  form nitrous
acid,  which  will  react with  some organic compounds  to  form oximes.    These

                                  9012A - 1                       Revision  1
                                                                  January 1995

-------
compounds once formed will decompose under test conditions to generate HCN.  The
possibility of interference of nitrate and nitrite is eliminated by pretreatment
with  sulfamic  acid  just before  distillation.    Nitrate  and  nitrite  are
interferences when present at levels higher than 10 mg/L  and in conjunction with
certain organic compounds.

      3.5    Thiocyanate is reported to be an  interference when present at very
high levels.  Levels of 10 mg/L were not found to interfere in Method 9010.

      3.6    Fatty acids,  detergents, surfactants, and other compounds may cause
foaming during the distillation when they are present in large concentrations and
will make  the  endpoint  of the  titration difficult  to detect.  They may be
extracted at pH 6-7.

4.0   APPARATUS AND MATERIALS

      4.1    Reflux distillation apparatus such as shown in Figure 1 or Figure
2.  The boiling flask should be of  one  liter size with inlet tube and provision
for  condenser.    The  gas  scrubber  may  be a  270-mL  Fisher-Milligan  scrubber
(Fisher, Part No.  07-513 or equivalent).   The  reflux apparatus may be a Wheaton
377160 distillation unit or equivalent.

      4.2    Automated continuous-flow analytical instrument with:

             4.2.1     Sampler.

             4.2.2     Manifold.

             4.2.3     Proportioning pump.

             4.2.4     Heating bath with distillation  coil.

             4.2.5     Distillation head.

             4.2.6     Colorimeter  equipped with  a 15-mm flowcell  and  570 nm
      filter.

             4.2.7     Recorder.

      4.3    Hot  plate stirrer/heating mantle.

      4.4    pH meter.

      4.5    Amber light.

      4.6    Vacuum  source.

      4.7    Refrigerator.

      4.8    5 mL microburette.


                                   9012A - 2                       Revision 1
                                                                  January  1995

-------
      4.9    7 Class A volumetric flasks - 100 and 250 ml.

      4.10   Erlenmeyer flask  - 500 ml.


5.0   REAGENTS

      5.1    Reagent  grade  chemicals  shall   be  used  in all  tests.   Unless
otherwise  indicated,  it  is  intended  that  all reagents  shall conform  to the
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where such  specifications  are available.   Other grades  may be used,
provided it is  first  ascertained that the reagent  is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2    Reagent water.   All  references  to water in this  method refer to
reagent water,  as defined in Chapter One.

      5.3    Reagents for sample collection,  preservation, and handling

             5.3.1     Sodium arsenite  (0.1N), NaAs02.  Dissolve 3.2 g NaAs02 in
      250 ml water.

             5.3.2     Ascorbic acid,  C6H806.

             5.3.3     Sodium  hydroxide  solution (50%),  NaOH.   Commercially
      available.

             5.3.4     Acetic  acid   (1.6M)  CH3COOH.     Dilute  one  part  of
      concentrated acetic acid with 9 parts of water.

             5.3.5     2,2,4-Trimethylpentane,  C8H18.

             5.3.6     Hexane,  C6H14.

             5.3.7     Chloroform,  CHC13.

      5.4    Reagents for cyanides amenable to chlorination

             5.4.1     Calcium hypochlorite solution (0.35M),  Ca(OCl)2.  Combine
      5 g of calcium hypochlorite and 100 ml  of water.   Shake before using.

             5.4.2     Sodium hydroxide solution  (1.25N),  NaOH.  Dissolve 50 g
      of NaOH in  1 liter of water.

             5.4.3     Sodium arsenite  (0.1N).  See Step  5.3.1.

             5.4.4     Potassium iodide starch paper.
                                   9012A  - 3                       Revision 1
                                                                  January 1995

-------
      5.5    Reagents for distillation

             5.5.1     Sodium hydroxide (1.25N).   See Step 5.4.2.

             5.5.2     Bismuth nitrate (0.062M),  Bi(NO)3 •  5H20.  Dissolve 30 g
      Bi(NO)3 • 5H20  in  100 ml of water.  While stirring, add 250 ml of glacial
      acetic acid, CH3COOH.   Stir  until  dissolved and dilute  to  1  liter with
      water.

             5.5.3     Sulfamic acid (0.4N),  H2NS03H.  Dissolve 40 g H2NS03H in
      1 liter of water.

             5.5.4     Sulfuric acid (18N), H2S04.  Slowly  and carefully add 500
      mL of concentrated H2S04 to 500  ml  of water.

             5.5.5     Magnesium chloride solution (2.5M), MgCl2» 6H20.  Dissolve
      510 g of MgCl2  • 6H20  in 1 liter of water.

             5.5.6     Lead acetate paper.

      5.6    Reagents for automated colorimetric determination

             5.6.1     Pyridine-barbituric   acid   reagent:     Place  15  g  of
      barbituric  acid  in  a 250-mL  volumetric  flask,  add  just  enough  reagent
      water to wash the  sides  of the flask,  and wet the barbituric acid.  Add 75
      ml of pyridine and mix.  Add 15 ml of concentrated HC1,  mix,  and cool to
      room temperature.   Dilute to 250  mL with  reagent water and mix.   This
      reagent is stable for approximately six months if stored in a cool, dark
      place.

             5.6.2     Chloramine-T solution:  Dissolve 2.0 g  of white,  water
      soluble  chloramine-T  in 500 mL  of reagent water and refrigerate  until
      ready to use.

             5.6.3     Sodium hydroxide,  1 N:   Dissolve 40  g of NaOH in reagent
      water, and dilute to 1 liter.

             5.6.4     All  working  standards  should  contain 2 mL of  1  N NaOH
      (Step 5.6.3) per 100 mL.

             5.6.5     Dilution water and receptacle wash water (NaOH, 0.25 N):
      Dissolve 10.0 g NaOH in 500 mL of reagent water.  Dilute to 1 liter.

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1    All samples must be collected using  a  sampling plan that addresses
the considerations discussed in Chapter Nine.

      6.2    Samples  should  be  collected in  plastic  or  glass  containers.  All
containers must be thoroughly cleaned and rinsed.

                                   9012A  - 4                       Revision 1
                                                                  January 1995

-------
      6.3    Oxidizing  agents  such as chlorine  decompose most cyanides.   To
determine whether oxidizing agents are present, test a drop of the sample with
potassium  iodide-starch test  paper.    A blue  color  indicates  the  need  for
treatment.  Add 0.1N sodium arsenite solution a few mL at a time until  a drop of
sample produces  no  color on the  indicator  paper.   Add an additional  5  ml of
sodium arsenite solution for each  liter of sample.  Ascorbic acid can be used as
an alternative although it is not  as effective  as arsenite.  Add a few crystals
of ascorbic  acid at a  time until a drop of  sample produces no  color  on  the
indicator paper.  Then add an additional  0.6 g of ascorbic acid for each liter
of sample volume.

      6.4    Aqueous samples must be  preserved by  adding 50% sodium hydroxide
until the pH is greater than or equal  to 12 at the time of collection.

      6.5    Samples should be chilled to 4°C.

      6.6    When properly  preserved, cyanide  samples  can be stored for up to
14 days prior to sample preparation steps.

      6.7    Solid and oily wastes may be extracted  prior  to analysis by Method
9013 (Cyanide Extraction Procedure for Solids  and Oils).   It uses a dilute NaOH
solution (pH = 12) as the extractant.   This yields extractable cyanide.

      6.8    If fatty acids, detergents,  and surfactants are a problem, they may
be extracted using the following procedure.   Acidify the sample with acetic acid
(1.6M) to pH 6.0 to 7.0.

             CAUTION:  This procedure  can produce lethal HCN  gas.

Extract with isooctane, hexane, or chloroform (preference in  order named) with
solvent volume equal to 20% of the sample  volume.   One extraction  is usually
adequate to reduce the compounds below the  interference level.  Avoid multiple
extractions or a long contact time at low pH in order to keep  the loss of HCN at
a minimum.  When the extraction is completed,  immediately raise the pH of the
sample to above 12 with 50% NaOH solution.

7.0   PROCEDURE

      7.1    Pretreatment for cyanides amenable to chlorination

             7.1.1     This test must be performed  under  amber  light.  K3[Fe-
      (CN)6]  may decompose  under   UV  light  and hence  will test  positive  for
      cyanide amenable  to  chlorination  if  exposed to  fluorescent  lighting or
      sunlight. Two  identical sample aliquots are required  to determine cyanides
      amenable to chlorination.

             7.1.2     To one 500 ml sample  or to a sample diluted to 500 ml, add
      calcium hypochlorite solution dropwise while agitating and maintaining the
      pH between  11 and  12 with   1.25N  sodium hydroxide  until  an excess  of
                                  9012A  - 5                       Revision 1
                                                                  January 1995

-------
chlorine is present  as  indicated  by Kl-starch paper turning  blue.   The
sample will be subjected to alkaline chlorination by this step.

       CAUTION: The initial reaction product of alkaline chlorination is
       the very toxic gas  cyanogen  chloride;  therefore,  it is necessary
       that this reaction be performed in a hood.

       7.1.3     Test for excess chlorine with Kl-starch paper and maintain
this excess for one hour with continuous agitation.  A distinct blue color
on the test paper indicates a  sufficient  chlorine  level.   If necessary,
add additional calcium hypochlorite solution.

       7.1.4     After one hour, add 1 ml portions of 0.1N sodium arsenite
until Kl-starch paper shows  no residual   chlorine.   Add 5  ml  of excess
sodium arsenite to ensure the presence of excess reducing agent.

       7.1.5     Test for total cyanide  as  described below  in  both  the
chlorinated and  the unchlorinated  samples.   The  difference  of  total
cyanide  in the  chlorinated  and  unchlorinated samples  is the  cyanide
amenable to chlorination.

7.2    Distillation Procedure

       7.2.1     Place 500 ml of sample, or sample diluted to 500 ml in the
one liter boiling flask.  Pipet 50 ml of  1.25N sodium hydroxide into the
gas scrubber.   If the apparatus in Figure 1  is used, add water until  the
spiral is covered.  Connect the boiling flask,  condenser, gas scrubber and
vacuum trap.

       7.2.2     Start a slow  stream of air entering the boiling flask by
adjusting the  vacuum source.  Adjust the vacuum so that approximately two
bubbles of air  per second enter the boiling flask through the air inlet
tube.

       7.2.3     If samples are known or suspected to contain sulfide, add
50 mL of 0.062M bismuth  nitrate solution through the air inlet tube.  Mix
for three  minutes.   Use lead  acetate paper to check the  sample for the
presence of sulfide.  A  positive test is indicated by a black color on the
paper.

       7.2.4     If samples are known or  suspected  to contain nitrate or
nitrite, or if bismuth nitrate was added to the sample, add 50 ml of 0.4N
sulfamic acid  solution through the air inlet  tube.  Mix for three minutes.

       Note:  Excessive  use of sulfamic acid could create method bias.

       7.2.5     Slowly  add 50  mL of 18N sulfuric acid  through the air
inlet tube.   Rinse the  tube with  water and  allow the airflow to mix the
flask contents for three minutes.   Add 20 ml  of 2.5M magnesium chloride
through the air inlet and wash the  inlet  tube with a stream of water.


                            9012A - 6                       Revision 1
                                                            January 1995

-------
        7.2.6     Heat the solution to boiling.  Reflux for one hour.  Turn
 off heat and continue the airflow for at least 15 minutes.  After cooling
 the  boiling  flask,  and  closing the vacuum  source,  disconnect  the gas
 scrubber.

        7.2.7     Transfer the  solution  from  the scrubber  into  a 250-mL
 volumetric flask.  Rinse the scrubber into the volumetric flask.  Dilute
 to volume with water.

 7.3    Automated colorimetric determination:

        7.3.1     Set up the manifold in a hood or a well-ventilated area
 as shown in Figure 3.

        7.3.2     Allow colorimeter and recorder to warm up for 30  min.  Run
 a baseline with  all  reagents,  feeding reagent water through  the sample
 line.

        7.3.3     Place appropriate  standards in  the  sampler  in  order of
 increasing concentration.   Complete  loading of the sampler tray  with
 unknown samples.

        7.3.4     When the baseline  becomes steady,  begin the  analysis.

 7.4    Standard curve for samples without sulfide

        7.4.1     Prepare  a  series   of  standards by  pipetting  suitable
 volumes  of working  standard  potassium  cyanide  solution  into  250-mL
 volumetric flasks.  To each flask, add 50 mL of 1.25N sodium hydroxide and
 dilute to  250  ml with water.   Prepare  using the  following table.   The
 sodium hydroxide concentration will  be 0.25N.

ml of Working Standard Solution        Concentration
       (1 mL = 10 uq  CN")                   (uq CNVL1
             0                            Blank
           1.0                             40
           2.0                             80
           5.0                            200
          10.0                            400
          15.0                            600
          20.0                            800

        7.4.2     After the standard solutions have been prepared according
 to the table above,  pipet 50 mL  of  each  standard solution into  a 100-mL
 volumetric flask and  proceed to Steps 7.3.2 and 7.3.3 to obtain absorbance
 values for the  standard  curve.  The final concentrations for the  standard
 curve  will  be  one  half  of  the  amounts  in  the  above  table  (final
 concentrations  ranging from 20 to 400 ug/L).
                             9012A  - 7                       Revision 1
                                                             January 1995

-------
             7.4.3     It is recommended that at least two standards  (a high and
      a low) be distilled and compared to  similar values  on the curve to ensure
      that the  distillation  technique is  reliable.   If  distilled standards do
      not agree within +  10% of the  undistilled  standards,  the analyst should
      find the cause of the apparent error before proceeding.

             7.4.4     Prepare a standard  curve ranging from 20 to 400 jug/L by
      plotting absorbance of standard versus the cyanide concentration

      7.5    Standard curve  for samples with sulfide

             7.5.1     It is imperative that all  standards be distilled in the
      same  manner as  the  samples  using  the  method  of standard  additions.
      Standards distilled by this  method will  give a  linear curve,  at  low
      concentrations,  but   as   the   concentration   increases,  the  recovery
      decreases.  It is recommended that at least five standards be distilled.

             7.5.2     Prepare a series of standards  similar  in  concentration to
      those mentioned  in Step  7.4.1  and  analyze as  in  Step 7.3.   Prepare a
      standard  curve by  plotting  absorbance of  standard versus  the  cyanide
      concentration.

      7.6    Calculation:  Prepare a standard curve by plotting peak heights of
standards against  their concentration values. Compute concentrations of samples
by comparing sample peak heights with the standard curve.

8.0   QUALITY CONTROL

      8.1    Refer to Chapter One for specific quality control procedures.

      8.2    Verify the calibration curve  with  an independent calibration check
standard.   If  the standards are not  within  15% of  the  expected  value,  a new
recalibration curve is required.  Verify the calibration curve with every sample
batch by analyzing a mid-range standard.

      8.3    Run  one matrix  spike  sample for every  10  samples to  check the
efficiency of sample distillation.  A matrix  spike should be prepared by adding
cyanide from the working standard or intermediate standard to 500 mL of sample
to ensure a concentration of approximately 40 /xg/L.   Both the matrix duplicate
and matrix spike duplicate are brought through the entire  sample preparation and
analytical process.

      8.4    The method of standard additions shall be used for the analysis of
all samples that suffer from matrix  interferences such  as samples which contain
sulfides.

9.0   METHOD PERFORMANCE

      9.1    Precision and  accuracy data  are not available at  this time.
                                   9012A -  8                      Revision 1
                                                                  January 1995

-------
10.0  REFERENCES

1.    Annual Book of ASTM Standards, Part 31,  "Water,"  Standard D2036-75, Method
B, p. 505 (1976).

2.    Goulden, P.O., B.K.  Afghan, and P. Brooksbank, Determination of Nanogram
Quantities of Simple  and  Complex Cyanides in Water,  Anal.  Chem.,  44(11), pp.
1845-49 (1972).

3.    Standard Methods for the Examination of Water and Wastewater, 14th ed., pp.
376 and 370, Method 413F and D (1975).

4.    Technicon AutoAnalyzer II Methodology,  Industrial Method No. 315-74 WCUV
Digestion and Distillation, Technicon Industrial Systems, Tarrytown, New York,
10591 (1974).
                                  9012A - 9                       Revision 1
                                                                  January 1995

-------
                               Connecting Tubing
      Allihn Condenser
    Air Inlet Tube
One-Liter
Boiling Flask
                                                   Suction
  Figure 1.  Apparatus for Cyanide Distillation
                     9012A -  10
Revision 1
January 1995

-------
 Cooling Water
Inlet Tube *
Screw Clamp
     I
      Heater •»
                                      To Low Vacuum Source
                                      Gas Scrubber
                          Condenser
                              Distilling Flask
                   O
 Figure  2.   Cyanide Distillation Apparatus
                9012A - 11
           Revision 1
           January 1995

-------
                                      a


                                      i
                                      o
                                      
-------
                                        METHOD 9012A

              TOTAL AND AMENABLE CYANIDE  fCOLQRIMETRIC.  AUTOMATED UV)
  7.1  Pretreat to
    determine
cyanides amenable
  to chlormation.
7.2.1  Place sample
  in flask; pipet
 sodium hydroxide
  into absorbing
      tube.
                               7.2.4
                            Are samples
                            suspected to
                             contain NO2
                               and/or
                                No3?
                             7.2.4 Add
                              sulfamic
                            acid solution
                            through air
                             inlet tube.
 7.2.2 Introduce
  air stream into
   boiling flask.
                         7.2.5  Add
                          rinse tube with
                           Type II water
                          add magnesium
                             chloride.
      7.2.3
     Positive
     sulfide
      test?
  7.2.3 Treat
   sample by
adding bismuth
nitrate solution.
   7.2.6 Boil
   solution;
  reflux; cool;
   close off
vacuum source.
                                                      7.2.7  Drain
                                                        solution
                                                     from absorber
                                                       into flask.
                                                      7.3 Perform
                                                        baseline
                                                      colorimetric
                                                       analysis.
                                        9012A -  13
                                                    Revision 1
                                                    January 1995

-------
                           METHOD 9012A

  TOTAL AND  AMENABLE CYANIDE  (COLORIMETRIC.  AUTOMATED UV)
                            (CONTINUED)
  7.5.1 Distill
  standards in
  same manner
   as sample.
  7.4
 Does
sample
contain
sulfide?
7.4.1 Prepare a
   series of
 CN standards.
 7.5.2 Prepare
standard curve
of absorbances.
                       7.4.2 Distill at
                         least two
                     standards to check
                         distillation
                        techniques.
                                                     7.4.3 Prepare
                                                    standard curve
                                                    of absorbances.
                          7.6 Compute
                         concentrations.
                        7.4.4 Check
                        efficiency of
                          sample
                         distillation.
                              9012A  -  14
                                     Revision 1
                                     January 1995

-------
                                  METHOD 9023

                  EXTRACTABLE ORGANIC  HALIDES  (EOX)  IN SOLIDS
1.0   SCOPE AND APPLICATION

      1.1   This method is to  be used for the determination  of total extractable
organic halides  (EOX)  as CV  in  solids.   EOX  is  defined as the  sum of those
organic halides which  are extracted  and detected  by pyrolysis/microcoulometry
under the  conditions specified in this method.    Extractable  organic halides
containing  chlorine,  bromine,  or  iodine  are  detected.   However,  fluorine
containing species are not detected by this method.

      1.2   This method has been evaluated for solid wastes,  soils,  and suspended
solids isolated from industrial wastewater.

      1.3   This method is recommended for use in the concentration range from
the MDL up to 1000 x MDL  (see Section 9.1).

      1.4   This method is restricted to  use by,  or under  the supervision of,
analysts experienced in the operation  of a pyrolysis microcoulometer and in the
interpretation of the results.

      1.5   Since this  method does not identify  individual components,  it is
advisable  that  compound  specific techniques  be  employed to determine  the
individual components  present in  samples exhibiting significant  EOX levels,
unless the nature of the  sample is already known.

2.0   SUMMARY OF METHOD

      2.1   A 1-gram aliquot of solid  sample is extracted with ethyl acetate by
sonification to  isolate organic  halides.   A 25 juL aliquot of  the extract is
either injected  or  delivered by boat  inlet  into  a pyrolysis  furnace  using  a
stream of C02/02 and the hydrogen halide (HX)  pyrolysis product is determined by
microcoulometric titration.

3.0   INTERFERENCES

      3.1   Method  interferences   may  be  caused   by  contaminants,  reagents,
glassware, and other sample processing hardware. All  of  these materials must be
routinely demonstrated to  be free from interferences under the conditions of the
analysis by running method blanks.

            3.1.1 Glassware must be scrupulously cleaned.  Clean all  glassware
      as soon as possible  after use by treating with chromate cleaning solution.
      This should be followed  by detergent washing in hot water.  Rinse with tap
      water and distilled water,  drain dry, and  heat  in a muffle furnace at
      400°C for 15  to  30  minutes.   Volumetric ware should not  be  heated  in  a
      muffle  furnace.    Glassware  should be  sealed and  stored  in  a  clean


                                   9023 - 1                       Revision 0
                                                                  January 1995

-------
      environment after drying and cooling to prevent any accumulation of dust
      or other contaminants.
            3.1.2       The  use  of high  purity reagents  and  gases  helps  to
      minimize interference problems.
            3.1.3 The use of non-TFE (polytetrafluoroethylene)  plastic tubing,
      non-TFE thread sealants, or flow controllers with rubber components in the
      purge gas stream should be avoided.
      3.2   Samples  can  be  contaminated   by  diffusion  of volatile  organics
(particularly solvents such as methylene chloride) through the septum seal  into
the sample during shipment and storage.
      3.3   All operations should be carried out  in  an area where halogenated
solvents, such as methylene chloride,  are  not being used.
      3.4   Certain  inorganic halide  salts (e.g.. mercuric chloride)  will  be
extracted, and therefore interfere to some extent.
4.0   APPARATUS AND MATERIALS
      4.1   Modified  Dohrmann  microcoulometric-titration  system  DX-20,  or
equivalent, containing the following components:
            4.1.1 Solvent injection system.
            4.1.2 Pyrolysis furnace.
            4.1.3 Titration cell.
      4.2   Boat inlet or Microsyringes -  10, 25  /nL  with 26 gauge 4-inch-long
      needle.
      4.3   Laboratory centrifuge to hold  15 mL conical centrifuge tubes.
      4.4   Sonic bath or sonic probe to  fit 10 mL vial.   A power level  of at
least 200 watts is required.
      4.5   Centrifuge Tubes - 15 mL,  conical,  with Teflon*-lined  screw caps.
      4.6   Vials - 10 mL, with Teflon*-!ined screw caps.
      4.7   Metal spatula
      4.8   Disposable Pasteur pipettes and bulbs.
                                   9023 - 2                       Revision 0
                                                                  January 1995

-------
5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests.  Unless otherwise
indicated, it is intended that all reagents  shall conform to  the specifications
of the Committee on Analytical Reagents of the American Chemical Society, where
such specifications  are  available.   Other grades may  be  used, provided it  is
first ascertained that the reagent is of sufficiently high  purity  to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent  water.   All references  to  water in this  method  refer  to
reagent water, as defined in  Chapter One.

      5.3   Carbon dioxide gas  (C02).  99.9 percent purity.

      5.4   Oxygen (02).   99.9 percent purity.

      5.5   Ethyl acetate (C4H802).   Pesticide quality  or equivalent.

      5.6   1,2,4-Trichlorobenzene (C6H3C13).  99 percent.

      5.7   Acetic acid (C2H402), 70% in water.   Dilute 7 volumes of acetic acid
with 3 volumes of water.

      5.8   Trichlorobenzene  solution (C6H3C13), stock  (1 jtiL  =  jug  Cl).  Prepare
a stock solution by  accurately  delivering  117 /iL (170 mg) of  trichlorobenzene
into a 100-mL volumetric flask  and dilute to volume with ethyl acetate.

      5.9   Trichlorobenzene  solution (C6H3C13),  calibration (1 /uL  =  100 ng Cl).
Dilute 10 ml of the trichlorobenzene stock solution to 100 ml with ethyl acetate.

      5.10  Sodium   chloride  (NaCl)  calibration  standard,   (1  /zg   CV/juL).
Accurately weigh 0.1648  g of sodium chloride into a 100-mL volumetric flask.
Dilute to volume with reagent water.

6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   All samples must  be collected using a sampling plan that addresses
the considerations discussed  in Chapter Nine.

      6.2   All samples must be  iced or refrigerated from the time  of collection
until analysis.
      6.3   All samples  should  be collected in bottles  (at  least 25 mL) with
Teflon* septa and  be protected from light.  If this is not possible, use amber
glass 250-mL bottles fitted with Teflon -lined caps.  Foil may be substituted for
Teflon* if the  sample is not corrosive.  Fill the sample  bottle  as  completely as
possible to minimize headspace until time of analysis.  Samples must be preserved
by acidification  to pH < 2 with  sulfuric acid, stored  at  4°C, and protected
against loss of volatiles by  eliminating headspace in the container.  Samples
                                   9023 - 3                       Revision 0
                                                                  January 1995

-------
should be analyzed within 28 days.   The  container must be washed and muffled at
400°C before use, to minimize contamination.

      6.4   If  the  analysis  is  to be  conducted  on  suspended solids  from a
wastewater sample, isolate the solids by centrifugation, weigh the wet solids,
and analyze immediately.  Determine the  dry weight of a  separate portion of the
wet solids by heating overnight at 110'C.

      6.5   All glassware must be dried  prior  to use according  to the protocols
discussed in Sec 3.1.1.

7.0   PROCEDURE

      7.1   Calibration

            7.1.1 Assemble  the  solvent  injection/pyrolysis/microcoulometric
      titration   apparatus   shown   in   accordance  with   the  manufacturer's
      specifications.  Adjust the C02 flow to 300 mL/minute and the 02 flow to
      100 mL/minute  using  the  auxiliary  flow controllers  (bypass  the  flow
      controllers).  The pyrolysis furnace should be set at 800 ± 10°C.  Attach
      the titration cell to the  pyrolysis tube outlet and fill with electrolyte
      (70% acetic acid).

            7.1.2 Turn  on  the   instrument   and  allow  the  gas  flows  and
      temperatures to stabilize.  When the background current of the titration
      cell has stabilized, the instrument is ready for use.

            7.1.3 Calibrate  the  microcoulometric   titration  system  for  Cl"
      detection by injecting various amounts of the sodium chloride calibration
      standards directly  into the  titration cell and  integrating the response
      using the  POX  integration mode.   The range  of sodium chloride amounts
      should cover the  range of expected sample concentrations and should always
      be  less  than 80 jug Cl".   Over the  range  1  -  80 jug Cl" the integrated
      response should  read within  2% or 0.05 jug (whichever  is larger)  of the
      quantity injected.   If  this  calibration requirement  is not met then the
      instrument  sensitivity  parameters should be  adjusted  according  to the
      manufacturer's specifications to achieve accurate response.

            7.1.4 Check  the  performance  of the entire analytical  system by
      delivering three  25-juL  aliquots of the trichlorobenzene  calibrate standard
      into  the  furnace at  a  rate  of  1 //L/second.    The  mean of  these three
      analyses  should  be 2.2 -  2.8 /xg  Cl  and the percent  relative standard
      deviation should be 5% or  less.   If these criteria are not met the system
      should be  checked  as described in the  instrument maintenance manual in
      order to isolate the problem.

            7.1.5 Perform a blank ethyl  acetate standard (25-juL) each day.  If
      the integrated response is greater than  0.1 jug Cl", then  the system should
      be checked for sources of contamination.
                                   9023 - 4                       Revision 0
                                                                  January 1995

-------
      7.2   Transfer a 1-gram aliquot of the  solid  sample to a 10 ml vial using
a metal spatula.   Add  1  ml of reagent water and 5 ml  of ethyl  acetate to the
sample and cap tightly.

      7.3   Shake  the  sample  vigorously  for  thirty seconds  and  then place the
vial in a  sonic  bath  filled with water to a level of  -1 inch,  or agitate the
suspension directly using a sonic probe, if available.   Sonify the sample for 15
minutes if using a sonic bath or 5 minutes if using a sonic probe.

      7.4   Allow the suspension to settle for  10 minutes and then transfer the
upper layer (ethyl  acetate)  to a 15-mL conical centrifuge tube.  Cap the tube and
centrifuge at approximately 1000 x g for five minutes.

      7.5   Transfer the  ethyl  acetate  layer to a clean 10 ml  vial,  cap, and
store refrigerated until  analyzed.

      7.6   For analysis, withdraw  a 5  to 25 /iL aliquot of the ethyl  acetate
into a microsyringe  having a 4-inch long  needle  or  a  boat inlet.   Place the
pyrolysis/microcoulometer system into the POX integration mode and immediately
pierce the septum  and  position the tip of the  microsyringe into  the furnace.
Deliver the  sample at a  rate of approximately 1  /iL/second and  withdraw the
needle when sample delivery is complete.

      7.7   After  the  10-minute  integration  cycle  is  complete  record  the
integrated  response.    If  the  response  exceeds  the  working  range  of  the
instrument, repeat the  analysis  after dilution of the extract with reagent grade
ethyl acetate.

      7.8   Determine the EOX concentration in the sample as follows:

      where:   EOX Concentration,  ug/g as  Cl~ = -^—I x 10(
                                                   Ws  x  VT

      Qs  = Quantity of  EOX as ;ug of Cl"  in  the aliquot  injected.

      V, = Volume of aliquot injected in /iL.

      VE = Total  volume of extract  in ml.

      Ws = Weight of sample extracted in grams.

      7.9    Report results  in micrograms per gram.  When duplicate  and spiked
samples are analyzed,  report all data obtained with the sample results.

      7.10  For samples  processed  as part of  a set where  the spiked  sample
recovery  falls outside  of  the historically derived control limits,  data for the
affected  parameters must  be labeled as suspect.

      7.11  If the aqueous portion of a water sample, from  which the suspended
solids are being  analyzed, is expected to contain high levels of organic halide,


                                   9023 - 5                       Revision 0
                                                                  January 1995

-------
a 1-mL aliquot of the centrifuged  sample  should  be  analyzed.   The solids data
must then be corrected using the following equation:

      where:        EOX (corrected) = EOXS - EOXW x -^
                                                        WD
      EOXS = EOX in  wet  solids,  /ug/g as Cl

      EOXW = EOX in water sample, pg/g as Cl

      Ws = Wet  weight of solids,  grams

      WD = Dry  weight of solids,  grams

8.0   QUALITY CONTROL

      8.1   Each laboratory  that uses this  method  is required to operate a formal
quality control program.  The minimum requirements of this program consist of an
initial demonstration of laboratory capability and the analysis  of spiked samples
as a continuing check on performance.  The  laboratory  is  required to maintain
performance records  to define the quality  of data that  is  generated.

      8.2   Before performing any  analyses,  the analyst must  demonstrate  the
ability to generate  acceptable accuracy and precision with this method.

            8.2.1 Select a trichlorobenzene spike concentration representative
      of the expected levels in the samples.  Using  stock  standards,  prepare a
      quality control check sample concentrate in ethyl  acetate 100 times more
      concentrated than  the selected concentration.

            8.2.2 Place  a minimum  of six  1-gram  aliquots  of an uncontaminated
      soil sample in 10  mL vials.  Spike  four of the samples with  10  /zL of the
      check sample,   cap  the vials, shake  vigorously,  and allow the  spike  to
      equilibrate with the sample by standing overnight.   Analyze  the aliquots
      according to the procedure beginning in Step 7.2.

            8.2.3 Calculate the average percent recovery, (R), and the standard
      deviation of the percent recovery (S), for the results.   Soil background
      corrections must be made before R and S calculations are performed.

            8.2.4   Acceptance limits for recovery and precision must be derived
      from repeated  analyses of the standard discussed  in  section  8.2.1.   Base
      the accuracy acceptance criteria on +/- 3 standard  deviations  from  the
      mean  recovery  and  the  precision acceptance  criteria  on  the  relative
      standard  deviation.  If the recovery for a particular parameter does  not
      fall  within  the  control   limits  for  method  performance,  the  results
      reported  for that  parameter in all  samples processed as  part of the same
      set must  be qualified as described  in Step 7.10.

      8.3   The laboratory must spike and analyze  a minimum of 10% of all samples
to monitor continuing laboratory performance.


                                   9023 -  6                       Revision  0
                                                                  January 1995

-------
      8.4   Each day,  the analyst must  demonstrate,  through  the  analysis of
uncontaminated soil,  that interferences  from the analytical  system are under
control.

9.0   METHOD PERFORMANCE

      9.1   The  method  detection  limit  (MDL),  is  defined   as  the  minimum
concentration  of a  substance  that  can  be measured  and  reported with  99%
confidence that the value is above zero.   An MDL of 10 jitg/g was obtained using
ethyl acetate standards.  The MDL actually achieved in a given analysis will vary
depending on instrument sensitivity and matrix effects.

      9.2   In a single laboratory, using  solid spiked at various levels,  the
average recoveries presented in Table 1 were obtained.

10.0  REFERENCES

      10.1  "Development and Evaluation of  Methods for  Total Organic Halide and
Purgeable Organic Halide in  Wastewater".  EPA-600/4-84-008,  PB84-134337 (NTIS).
U.S.  Environmental  Protection  Agency,   Environmental  Monitoring and  Support
Laboratory - Cincinnati, Ohio 45268,  January 1984.
                                   9023 - 7                       Revision 0
                                                                  January 1995

-------


















LU
_l
ea-
se
s>
*/7
0
_j
o
co
CO
o
>
0
1 1
*^m
<
£
0
111
UJ
u
*^B
§
O
1 1
U—
a:
LU
bLJ
o.
a
o
1
^k


*
^^
Li_l
UJ
j
wJ
23
^^
r"

























S-
0)
JD
£ ^4—
3 O
Z

C
-a o
i- ••-
("0 +J
~O ftf
c ••-
+j CU
OO Q
CU •»->
C7) C
rd CU
s_ u
CU S-
> 0)
< ex



CU
CU
— '
CU
.,_
o
00












CU
V
•^
OL
OO






















CO
CU
+J
fO
CJ
•r-
^^
ex
CU
a:
s-
a>
>
o
u
CU
ex:
(4-
0
i-
CU
>
o
u
CU
°£



to

\
5*


w
O)
3.








T3
c
3
0
a.
E
O
O









CU

a.
E
to
oo











co








i— i
i— i





CT>
l£>





VO





O
i — i




CU
c
N
C
CU
JD
0
s-
o
^~
^
u
•1—
s-
t—
1
*»•
csT
^





E
r—
•^
O
oo

c
rC
CU

O
E





CO CO








o in
CM •— 1





CO 00
o in
»— i





en 10
CM CM





O% P1"*
^* ^^




CU CU
c c
CU -r-
N t—
C -i-
cu c

o o
t- S-
0 O
^— ^^
g^ _^
u u
,f~ .^
S_ i-
1— H-
1 1
^- u>
CM ^~
•— 1 CM





= s
^_ p~m
•t— *r—
0 0
OO 00

c c
fO (9
CU CU

o o
E r





CO








1— 1
I— 1





CO






in
CM





10
^^





o
c
CU
^
ex
0
S-
0
^^
^r
o
•I—
s-
h-
|
IO
•*"
CM





s
^_
'i—
O
00

c
(0
0)
r— •
o
E





co








co






^^
o
1— I





o
o
in




o
in
00



CU
c
CU
N
C
CU
.0
0

0
r*"
-C
u
•1—
s_
1—
1
*
CM
~T

(/) ^-«
T3 ^5
•r— -r—
i— U
O +*^
CO 00
3
"O ^3 4-^
CU C C
~O •— • CU
C 3
CU E.—
ex o <4-
to i- ^-
3 t- LU
OO





CO








00






CO
00





o
co





1— 1
in




CU
c
CU
rsj
c:
CU

0

0
("—
f~
y
•p—
s-
1—
1
Tt-
CM
i— l



^_
fm^.
•r—
CU (4-
+•* ^o
00 C
flS nJ
3 —1

T3 E
•i- o
^— ^
0 <4-
oo

























































E
0
i- CU
*4- C
CU
r CU r-
•o -»-> >i
eu >r— i—
S- 0 J=
Q 00 O
E
i o ir>




co








o
t— 1





^^
en





o
r-.
r— 1




O
cr>
CM



CU
c
CU
ISJ
C
CU
JO
O
s-
o
r—
J^
(J
•r-
s_
i—
i
^
CM
i— l



O)
C
•^
S-
3
^j
U
m
(4_
3
C
ro
3^
en
c en
O (""H
•r"
 >»
••- S-

CU 3
rv* ^




















00
1
CO
CM
O
en








































-------
                   METHOD 9023

EXTRACTABLE ORGANIC  HALIDES  (EOX)  IN SOLIDS
^
'
7.1.1 - 7.1.2 Assemble
and adjust gas flow
and allow temp, and
gas flow to stabilize.
>
r
7.1.3 - 7.1.5 Calibrate,
check the performance
and run standard on
the instrument.
1
f
7.2 Dissolve 1 g of solid
sample in reagent with
ethyl acetate solution.

                    7.3 Shake and
                   sonicate mixture.
                7.4 Transfer upper level
                   to and centrifuge.
                 7.5 Transfer the ethyl
                 acetate layer and store
                      for analysis.
                 7.6 - 7.7 Analyze sample
                   and run 10 minute
                    integration cycle.

                  7.8 Determine EOX
                concentration and report.
                                                           Revision 0
                                                           January 1995

-------
                                  METHOD  9057

           DETERMINATION OF CHLORIDE FROM HC1/C1, EMISSION SAMPLING
             TRAIN  (METHODS  0050  AND  0051) BY ANION CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   This method describes the analytical protocol for determination of
hydrogen chloride  (HC1,  CAS Registry Number 7647-01-0)  and  chloride (C12 CAS
Registry Number 7782-50-5) in stack gas emission samples collected from hazardous
waste and municipal waste incinerators using the midget impinger HC1/C12 sampling
train (Method 0051) or the isokinetic sampling train (Method 0050).

      1.2   The lower  detection  limit is 0.1  /zg of  chloride  (CV)  per  mL of
sample solution.  Samples with concentrations which exceed the linear range of
the analytical instrumentation may be diluted.

      1.3   This method is recommended for use only by analysts experienced in
the use of ion chromatography and in the interpretation of ion chromatograms.

2.0   SUMMARY OF METHOD

      2.1   The stoichiometry of HC1 and HC12 collection in the sampling train
(see Methods  0050  and 0051)  is  as  follows:  In the  acidified  water absorbing
solution, The HC1  gas is  solubilized and  forms chloride ions (C1-) according to
the following formula:

                             HC1  + H20 =  H30+ + CV

The C12 gas present in the emissions  has a very low solubility in acidified water
and  passes  through  to  the  alkaline  absorbing solution  where   it  undergoes
hydrolysis to form a proton  (H+), Cl" ,  and hypochlorous  acid  (HC10)  as  shown:

                          H20 +  C12 = H+ +CV +  HC10

Non-suppressed or suppressed ion  chromatography (1C) is used for analysis of the
CT.

3.0   INTERFERENCES

      3.1   Volatile  materials  which produce  chloride  ions  upon dissolution
during sampling are obvious  interferences in the measurement of HC1.  One likely
interferant is diatomic  chlorine (C12) gas which disproportionates to HC1 and
hypochlorous  acid  (HOC1)  upon dissolution in  water.   C12 gas exhibits  a low
solubility in water,  however,  and the use of acidic  rather than  neutral or basic
solutions for collection of hydrogen  chloride gas greatly reduces the dissolution
of any chlorine present.  Sampling a 400 ppm HC1 gas stream containing 50 ppm C12
with this method does not cause a significant bias. Sampling a 220 ppm HC1 gas
stream containing 180 ppm C12 results in a positive bias of 3.4 percent in the
HC1 measurement.  Other interferants have not been  encountered.
                                   9057 - 1                       Revision 0
                                                                  January 1995

-------
      3.2   Reducing agents  such  as S02 may cause a positive  bias  in the C12
measurement by the following reaction:

                           HC10 + HS03- = H2S04 + CV

4.0   APPARATUS AND MATERIALS

      4.1   Volumetric Flasks.  Class A, various sizes, as appropriate.

      4.2   Volumetric  Pipettes.  Class  A,  assortment,  to  dilute  samples  to
calibration range of the 1C.

      4.3   Ion Chromatograph. Suppressed or non-suppressed,  with a conductivity
detector  and  electronic integrator  operating in the  peak  area mode.   Other
detectors, a strip chart recorder, and peak heights may be used provided the 5
percent repeatability criteria for sample analysis and the linearity criteria for
the calibration curve can be met.

5.0   REAGENTS

      5.1   Reagent grade chemicals  shall be used in all tests.  Unless otherwise
indicated, it is intended that all reagents  shall conform to the specifications
of the Committee on Analytical Reagents  of the American Chemical Society, where
such specifications  are  available.   Other grades may  be  used,  provided it is
first ascertained that the reagent is of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent  water.    All  references to water  in the method  refer to
reagent water as specified by definition in Chapter One.

      5.3   Sulfuric acid (0.1 N), H2S04.  To prepare  100 ml, slowly  add 0.28 ml
of concentrated  H2S04 to  about  90 ml of water  while  stirring,  and  adjust the
final volume to 100 ml using additional  water.  Shake well to mix the solution.

      5.4   Sodium hydroxide  (0.1 N), NaOH.   To prepare 100 ml, dissolve 0.40 g
of solid NaOH in about 90 ml of water and adjust the final volume to 100 ml using
additional water.  Shake well to mix the solution.

      5.5   Reagent blank solutions. A separate blank solution  of each sampling
train regent used and collected in the field (0.1 N H2S04 and 0.1 N NaOH) should
be prepared  for analysis with  the  field samples.   For  midget impinger train
sample analysis, dilute 30 ml of each reagent with  rinse  water collected in the
as a  blank to the final  volume of  the  samples.   For  isokinetic  train sample
analysis, dilute 200 ml to the same  final volume as the field samples also using
the blank  sample of  rinse water.

      5.6   Sodium chloride, NaCl, stock standard solution. Solutions containing
a nominal  certified  concentration of 1000 mg/L  NaCl are commercially available
as convenient stock  solutions from which  working standards  can  be  made by
appropriate volumetric dilution.  Alternately, concentrated  stock solutions may
be produced from reagent grade NaCl  that  has been dried at 110'C for  two or more
hours and  then  cooled to room  temperature  in a desiccator  immediately before
weighing.  Accurately weigh  1.6 to  1.7 g of the dried  NaCl to within 0.0001 g,


                                    9057  - 2                       Revision 0
                                                                  January 1995

-------
dissolve in water, and  dilute  to  1  liter.   The exact Cl"  concentration can be
calculated using the equation:

                  M9 Cr/mL = g of NaCl x 1000 x 35.453/58.44

Refrigerate the stock standard solutions and store no longer than one month.

      5.7   Chromatographic eluent.   Effective eluent for  non-suppressed  ion
chromatography using a resin- or silica-based weak ion exchange column are a 4
mM potassium  hydrogen phthalate  solution,  adjusted  to  a pH  of 4.0  using  a
saturated sodium borate  solution, and a mM 4-hydroxy benzoate solution, adjusted
to a pH of 8.6 using 1 N sodium hydroxide.   An effective eluent for suppressed
ion chromatography is a solution containing 3 mM sodium bicarbonate and 2.4 mM
sodium carbonate.  Other dilute solutions buffered to  a similar pH that contain
no ions interfering with the Chromatographic  analysis may be  used.   If, using
suppressed ion chromatography,  the "water dip" resulting from sample injection
is interfering with the  chloride peak, use a 2 mM sodium hydroxide/2.4 mM sodium
bicarbonate eluent.

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING.

      6.1   Sample collection  using  the midget impinger  HC1/C12 train  or  the
isokinetic HC1/C12 train is described in Method 0051  or  0050,  respectively.

      6.2   Samples  should  be  stored   in   clearly   labeled,  tightly  sealed
containers between sample recovery and analysis.  They may be analyzed up  to four
weeks after collection.

7.0   PROCEDURE

      7.1   Sample preparation for  analysis.   Check  the  liquid  level  in each
sample, and determine if any sample was lost during shipment.   If a noticeable
amount of leakage has occurred, the volume can be determined from the difference
between the initial  and final  solution levels,  and this  value can  be used to
correct  the   analytical   results.     For   midget   impinger   train  samples,
quantitatively transfer  each sample  solution  to a  100-mL  volumetric flask and
dilute  to  100  mL  with  water.     For  isokinetic   sampling  train  samples,
quantitatively transfer each sample  to a volumetric flask  or graduated cylinder
and dilute with water to a final volume appropriate for all samples.

      7.2   Calibration of Ion Chromatograph.

            7.2.1 The ion Chromatographic conditions  will  depend on  the  type of
      analytical  column used  and whether  suppressed  or non-suppressed  ion
      chromatography  is used.    Prior to  calibration and  sample  analysis,
      establish  a stable  baseline.    Next,  inject   a  sample  of   water,  and
      determine if any CV appears  in the chromatogram. If  Cl" is present, repeat
      the load/injection procedure until no Cl" is present.

            7.2.2 To prepare the  calibration standards,  dilute  given amounts
      (1.0 mL or greater) of the  stock standard solution to convenient volumes,
      using 0.1  N  H2S04  or 0.1  N  NaOH as appropriate.  Prepare  at  least four
      standards that are within the linear range of the field samples.  Inject


                                   9057 - 3                       Revision 0
                                                                  January 1995

-------
      the calibration standards, starting with the lowest concentration standard
      first,  both  before  and  after  injection  of  the quality  control  check
      sample, reagent blank, and  field  samples.   This allows compensation for
      any instrument drift occurring during sample analysis.

            7.2.3 Determine the peak  areas,  or heights,  of the standards and
      plot individual values versus CV concentrations  in jug/mL.  Draw a smooth
      curve through the  points.   Use  linear  regression to calculate a formula
      describing the resulting linear curve.

      7.3   Sample analysis.   Between  injections  of the series of calibration
standards,  inject,  in duplicate,  the reagent blanks and  the  field samples,
including a matrix  spike  sample.   Measure the areas or heights (same  as done for
the  calibration  standards) of the CV peaks.    Use  the  average  response to
determine the concentrations  of the field samples,  matrix  spike,  and reagent
blanks using  the linear calibration  curve,   the results for  a reagent blank
should not exceed 10 percent of the corresponding value for  a field  sample.

      7.4   Calculations.  Retain  at least one extra decimal  figure beyond those
contained in the  available  data in intermediate calculations, and round off only
the final answer appropriately.

            7.4.1 Total /j,g HC1 per  sample.  Calculate as  described  below:

                  mHci =  (S'B)  x vs x 36.46/35.453                 (1)

            where:

                  mHd   =  Mass of HC1 in sample,  y^g,
                  S            = Analysis of sample, jug Cl'/mL,
                  B            = Analysis of reagent blank,  /ug  CV/mL,
                  Vs    =  Volume  of filtered and  diluted  sample, ml
                  36.46 =  Molecular weight of HC1, /ig/^g-mole,
            and
                  35.45 =  Atomic  weight of CV, jug//Ltg-mole.

            7.4.2 Total nq C12 per  sample.  Calculate  as  described  below:

                 mcl2 =  (S-B) x  V2  x  70.91/35.45              (2)

            where:

                  mcl2    =  Mass of C12 in sample,   /zg
                  70.91  =  Molecular weight of C12 /ig/jug-mole,
            and
                  35.45  =  Atomic  weight of CV,  /Ltg//ig-mole.


            7.4.3 Concentration of  HC1  in the flue  gas: Calculate as described
      below:
                                C = K x m/V
                                           m(std)
                                    9057  -  4                        Revision 0
                                                                   January 1995

-------
            where:

                  C     = Concentration of HC1 or C12,  dry basis,  mg/dscm,
                  K     = 10"3 mg/jug,
                  m     = Mass of HC1 or C12 in sample,  ng
            and
                  Vm(stdi  =  Dry  9as  volume  measured  by  the  dry gas  meter,
                  corrected to standard conditions, dscm  (from Method 0050 or
                  Method 0051).

8.0   QUALITY CONTROL

      8.1   At the present time, a validated audit material does not exist for
the method.  However, it  is  strongly recommended that  a quality control check
sample and a matrix spike sample be used.

            8.1.1 Quality control check sample.   Chloride  solutions of reliably
      known  concentrations  are  available   for   purchase from  the  National
      Institute  of  Science and  Technology  (SRM  3182).   The  QC  check sample
      should be prepared in the appropriate absorbing reagent  at a concentration
      approximately equal to  the  mid-range  calibration  standard,   The quality
      control check sample should be injected, in duplicate,  immediately after
      the calibration standards have been  injected for  the first time.  The CV
      value  obtained  for  the check  sample  using the final  calibration curve
      should be within  10 percent of the known value for  the check sample.

            8.1.2 Matrix spike sample.  A portion of at least  one field sample
      should be used to  prepare a matrix spike sample.  Spike  the sample aliquot
      in the  range  of  the expected  concentration.   Analyze  the  matrix spike
      sample in  duplicate  along  with the  field  samples.   Based  on the matrix
      spike  results,  determine the  recovery for the  spiked  material.   This
      should be within  10 percent of the known spike value.

      8.2   Refer to Chapter One for additional quality control criteria.

9.0   METHOD PERFORMANCE

      9.1   The lower detection limit of the analytical  method is  0.1  jug of Cl"
per mL of sample solution.  Samples with concentrations  which exceed the linear
range of the 1C may be  diluted.

      9.2   The precision and bias of for  analysis of HC1  using this analytical
protocol have been measured in combination  with the midget  impinger HC1/C12 train
(Method 0051)  for sample  collection.   The within-laboratory relative standard
deviation is 6.2  percent and 3.2 percent at HC1 concentrations of 3.9 and  15.3
ppm, respectively.  The method does not exhibit any bias  for HCl when sampling
at C12 concentrations less than 50 ppm.

10.0  REFERENCES

1.    Steinsberger, S.C. and  J. H. Margeson, "Laboratory and Field  Evaluation of
      a  Methodology  for  Determination of  Hydrogen  Chloride  Emissions   from
      Municipal and Hazardous  Waste Incinerators," U.S. Environmental Protection


                                   9057 - 5                       Revision 0
                                                                  January  1995

-------
      Agency, Office of Research and Development, Report No. EPA 600/3-89/064,
      NTIS PB89 220586-AS.

2.    State of California, Air  Resources  Board,  Method  421, "Determination of
      Hydrochloric Acid Emissions from Stationary Sources" March 18, 1987.

3.    Entropy Environmentalists, Inc.,  "Laboratory  Evaluation of a Sampling and
      Analysis Method for Hydrogen Chloride emissions from Stationary Sources:
      Interim Report,"  EPA  Contract No.  68-02-4442, Research  Triangle Park,
      North Carolina, January 22, 1988.
                                    9057  -  6                        Revision  0
                                                                   January 1995

-------
                        METHOD 9057
DETERMINATION OF CHLORIDE FROM  HC1/C1, EMISSION SAMPLING
  TRAIN (METHODS 0050  AND 0051)  BY ANION  CHROMATOGRAPHY
                       7.2.1 Determine
                        1C conditions.
                            I
                        7.2.2 Prepare
                          calibration
                          standards.
                       7.2.3 Dtermine
                        peak areas of
                         standards.
                     7.3 Conduct sample
                      analysis with QC
                        and blanks.
                            I
                        7.4 Perform
                        calculations.
                          9057 - 7
Revision  0
January  1995

-------
                                  METHOD 9210

                    POTENTIOMETRIC DETERMINATION OF NITRATE
                IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
1.0   SCOPE AND APPLICATION

      1.1   This method can be used for measuring total solubilized nitrate in
drinking waters, natural surface waters, groundwaters, domestic and industrial
wastewaters, and in soil extracts (ASTM methods D4646-87, D5233-92 or D3987-
85).

            NOTE: This method is for the analysis of simple nitrate ion rather
            than total nitrate, as analysis using the ion-selective electrode
            is not preceded by a distillation step.

      1.2   The method detection limit is 2.0 mg/L.  Nitrate concentrations
from 0.2 to 1,000 mg/L may be measured.  However, using a linear calibration,
results less than 2 mg/L may be biased up to approximately 420 percent high;
results greater than 400 mg/L may be biased up to approximately 50 percent
low.

      1.3   ISEs must be used carefully, and results must be interpreted
cautiously, since an ISE may be affected by numerous analytical interferences
which may either increase or decrease the apparent analyte concentration, or
which may damage the ISE.  Effects of most interferences can be minimized or
eliminated by adding appropriate chemical reagents to the sample.  Obtaining
the most accurate results, therefore, requires some knowledge of the sample
composition.

            NOTE: ISE manufacturers usually include a list of interferences in
            the instruction manual accompanying an ISE, along with recommended
            methods for minimizing or eliminating effects of these
            interferences.

2.0   SUMMARY OF METHOD

      2.1   Total solubilized nitrate is determined potentiometrically using a
nitrate ion-selective electrode (ISE) in conjunction with a double-junction
reference electrode and a pH meter with an expanded millivolt scale or an ISE
meter capable of being calibrated directly in terms of nitrate concentration.

      2.2   Standards and samples are mixed 50:1 with an ionic strength
adjustment solution (ISA).  Calibration is performed by analyzing a series of
standards and plotting mV vs.  nitrate-nitrogen concentration on semilog paper
or by calibrating the ion meter directly in terms of nitrate concentration.

3.0   INTERFERENCES

      3.1   The nitrate electrode responds to numerous interfering anions.
Most of the interferants, however, can be rendered harmless by adding suitable
reagents.  Cyanide,  bisulfide, bicarbonate, carbonate, and phosphate are
removed by adjusting the solution pH to 4 with boric acid.  Chloride,  bromide,

                                   9210 - 1                       Revision 0
                                                                  January 1995

-------
and iodide are removed by adding silver sulfate solution.  Nitrite is also an
interferant, as shown in Table 1; nitrite is removed by adding sulfamic acid.
The amounts of silver sulfate and sulfamic acid needed will vary based on the
concentrations of interferants.  As a general guide, 1 ml of silver sulfate
will eliminate chloride interference in a 50 ml sample containing 35 mg/L Cl";
1 mL of sulfamic acid solution will eliminate nitrite interference in a 50 ml
sample containing 95 mg/L N02".

      3.2   Temperature changes affect electrode potentials.  Using an ISE
calibrated at 22°C, a 20.0 mg/L nitrate solution was measured as 20.6 mg/L at
22 "C and 12.9 mg/L at 32°C (see Ref. 4).  Therefore, standards and samples
must be equilibrated at the same temperature (+ 1°C).

      3.3   The user should be aware of the potential of interferences from
colloidal substances and that, if necessary, the samples may be filtered.

4.0   APPARATUS AND MATERIALS

      4.1   pH/mV meter capable of reading to 0.1 mV or an ISE meter.

      4.2   Nitrate ISE (Orion 9307 or equivalent) and double-junction
reference electrode (Orion 9002 or equivalent).

      4.3   Thermally isolated magnetic stirrer, Teflon®-coated stir bar, and
stopwatch.

      4.4   Volumetric flask, 100 mL.

5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests.  Unless
otherwise indicated, it is intended that all reagents shall conform to the
specifications of the Committee on Analytical Reagents of the American
Chemical Society, where such specifications are available.  Other grades may
be used, provided it is first ascertained that the reagent is of sufficiently
high purity to permit its use without lessening the accuracy of the
determination.

      5.2   Reagent water.  All references to water in this method refer to
reagent water, as defined in Chapter One.

      5.3   ISA adjuster solution (2M, (NH4)2S04):  Dissolve 26.4 g of
ammonium sulfate in reagent water to make 100 mL of solution.

      5.4   Boric acid (1M, H3B03):   Dissolve 6.2  g  of boric acid  in  reagent
water to make 100 mL of preservative solution (for numerous anions and
bacteria).

      5.5   Silver sulfate (0.05 M, Ag2S04)  to remove interferences  noted in
Step 3.1.  A saturated silver sulfate solution contains approximately 5.5 g/L
of solubilized silver.
                                   9210 - 2                       Revision 0
                                                                  January 1995

-------
      5.6   Sulfamic acid (0.1 M, HOS02NH2)  to  remove  nitrite  from sample,  as
noted in Step 3.1.

      5.7   Nitrate calibration stock solution (1,000 mg/L, N03"):  Dissolve
0.1631 g of potassium nitrate (dried two hours at 110'C and stored in a
desiccator) in reagent water, add 1.00 ml of preservative solution, and dilute
to 100 ml in a volumetric flask.  Store in a clean bottle.

      5.8   Nitrate calibration standards:  Prepare a series of calibration
standards by diluting the 1,000 mg/L nitrate standard.  A suitable series is
given in the table below.

            ml of 1,000 mg/L            Concentration when Diluted
             N03- Solution             to 50.0 mL  (mg/L N03'-N)
                0.0500                           1.00
                0.150                            3.00
                0.500                           10.0
                1.50                            30.0
                5.00                           100.0

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All samples must have been collected using a sampling plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   In most environmental samples, nitrate is not affected by
complexation, precipitation, inorganic oxidation-reduction reactions, and
protonation.  In the presence of a reducing agent (e.g., organic matter),
however, bacteria will utilize nitrate as an oxidant, causing a slow decrease
in the nitrate concentration.  This potential interference can be obviated by
using a preservative.  Therefore, samples must be preserved by adding 1 mL of
1M boric acid solution per 100 mL of sample.

      6.3   Samples should be stored at 4° C and  must be analyzed within  three
(3) days of collection.

7.0   PROCEDURE

      7.1   Calibration

            7.1.1 When using a nitrate ISE and a separate double-junction
      reference electrode, ensure that reference electrode inner and outer
      chambers are filled with solutions recommended by the manufacturer.
      Equilibrate the electrodes for at least one hour in a 100 mg/L nitrate
      standard before use.

            7.1.2 Calibrate the nitrate ISE using standards that narrowly
      bracket the expected sample concentration.   If the sample concentration
      is unknown, calibrate with 3.00 mg/L and 30.0 mg/L nitrate standards.
      Add 50.0 mL of standard, 0.50 mL of preservative solution, and 1.00 mL
      of ISA to a 100-mL beaker.  Add a Teflon®-coated magnetic stir bar,


                                   9210 - 3                       Revision 0
                                                                  January 1995

-------
      place the beaker on a magnetic stir plate, and stir at slow speed (no
      visible vortex).  Immerse the electrode tips to just above the rotating
      stir bar.  If using an ISE meter, calibrate the meter in terms of
      nitrate concentration following the manufacturer's instructions.  If
      using a pH/mV meter, record the meter reading (mV) as soon as the
      reading is stable, but in no case should the time exceed five minutes
      after immersing the electrode tips.  Prepare a calibration curve by
      plotting measured potential (mV) as a function of the logarithm of
      nitrate concentration.  The slope must be 54-60 mV per decade of nitrate
      concentration.  If the slope is not acceptable, the ISE may not be
      working properly.  For corrective action, consult the ISE operating
      manual.

      7.2   Allow samples and standards to equilibrate to room temperature.

      7.3   Prior to and between analyses, rinse the electrodes thoroughly
with reagent water and gently shake off excess water.  Low-level measurements
are faster if the electrode tips are first immersed five minutes in reagent
water.

      7.4   Add 50.0 ml of sample and 1.00 ml of ISA to a 100-mL beaker.  Add
a Teflon®-coated magnetic stir bar.  Place the beaker on a magnetic stir plate
and stir at a slow speed (no visible vortex).  Immerse the electrode tips to
just above the rotating stir bar.  Record the meter reading (mV or
concentration) as soon as the reading is stable, but in no case should the
time exceed five minutes after immersing the electrode tips.  If reading mV,
determine nitrate-nitrogen concentration from the calibration curve.

      7.5   When analyses have been completed, rinse the electrodes thoroughly
and store them in a 100 mg/L nitrate standard solution.  If the electrodes
will not be used more than one day, drain the reference electrode internal
filling solutions, rinse with reagent water, and store dry.

8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality control procedures.

      8.2   Initial Calibration Verification standard (ICV):  After performing
the calibration step (7.1), verify calibration by analyzing an ICV.  The ICV
contains a known nitrate concentration at the mid-range of the calibration
standards and is from an independent source.  ICV recovery must be 90-110
percent.  If not, the source of error must be found and corrected.  An
acceptable ICV must be analyzed prior to sample analysis.  The ICV also serves
as a laboratory control sample.

      8.3   Continuing Calibration Verification standard (CCV):  After every
10 samples, and after the final sample, a CCV must be analyzed.  The CCV
contains a known nitrate concentration at mid-calibration range.  CCV recovery
must be 90-110 percent.  If not, the error source must be found and corrected.
If ISE calibration has changed, all samples analyzed since the last acceptable
CCV must be re-analyzed.

      8.4   Reagent blank:  After the ICV and after every CCV, a reagent blank
must be analyzed.  A reagent blank is a 1 percent solution of preservative

                                   9210 - 4                       Revision 0
                                                                  January 1995

-------
solution in reagent water, mixed 50:1 with ISA.  The indicated reagent blank
concentration must be less than 1 mg/L nitrate.  If not, the contamination
source must be found and corrected.  All samples analyzed since the last
acceptable reagent blank must be re-analyzed.

      8.5   Matrix spike:  Follow the matrix spike protocols presented in
Chapter One.  The spike concentration must be 10 times the detection limit and
the volume added must be negligible (less than or equal to one-thousandth the
sample aliquot volume).  Spike recovery must be 75-125 percent.  If not,
samples must be analyzed by the method of standard additions.

9.0   METHOD PERFORMANCE

      9.1   In a single-laboratory evaluation, a series of standards with
known nitrate concentrations was analyzed with a nitrate ISE.  Measurements
were obtained over three consecutive days using an Orion 9307 nitrate ISE and
an Orion 9002 double-junction reference electrode connected to an Orion 940
ISE meter.  A two-point calibration (5.00 and 50.0 mg/L nitrate) was performed
prior to analysis.  The results are listed in Table 2.

      9.2   In the same study, three groundwater samples were spiked with
nitrate at four different concentrations and measured with the nitrate ISE.
(The groundwater samples initially contained <0.1-2.3 mg/L nitrate.)  Each
spiked sample was analyzed at each concentration, and the mean recoveries and
RSDs are given in Table 3.

      9.3   A 50 g portion of soil, which initially contained 0.7 mg/kg
nitrate, was spiked with 25.0 mg/kg nitrate to obtain an anion concentration
in a single extract volume within the linear range of the ISE.  The extract
was then analyzed for nitrate using this ISE method, and 89% of the soil spike
was recovered.

10.0  REFERENCES

      1.    Franson,  Mary Ann H., Ed.   Standard Methods for the Examination of
Water and Wastewater, 18th Edition.  American Public Health Association,
Washington, DC, 1992.

      2.    Model 93-07 Nitrate Electrode Instruction Manual.  Orion Research,
Inc.,  Boston,  MA, 1986.

      3.    Miller, E.L., Waltman, D.W., and Hillman, D.C.   Single-Laboratory
Evaluation of Fluoride, Chloride, Bromide,  Cyanide, and Nitrate Ion-Selective
Electrodes for Use in SW-846 Methods.   Lockheed Engineering and Sciences
Company for Environmental Monitoring Systems Laboratory, U.S. EPA.   September
1990.   EPA/600/X-90/221.
                                   9210 - 5                       Revision 0
                                                                  January 1995

-------
        Table 1.  Nitrate ISE Interferences
Analyte
Concentration
(mg/L)
25.0
25.0
25.0
25.0
Interference
None
0.01 M H2S04
100 mg/L N02'
100 mg/L N02' +
500 mg/L HOS02NH2
Measured
Concentration
(mg/L)
26
24.5
46
26
RSD
(%)
6.2
5.9
9.1
6.3
Table 2.  Results from a Single-Laboratory Accuracy
            Evaluation of a Nitrate ISE
Nitrate
Concentration
(mg/L)
0.100
0.200
0.500
1.00
2.00
5.00
10.0
20.0
50.0
100
200
400
1,000
Nitrate
Detected
(mg/L)
1.01
1.04
1.23
1.71
2.45
5.0
11.0
18.9
50
96
164
310
480
Nitrate
Recovery
(percent)
1,010
520
246
171
123
100
110
95
100
96
82
77
48
Rel. Std.
Deviation
(percent)
53
17
8
2
7
0
8
14
1
13
3
8
17
                      9210  -  6
Revision 0
January 1995

-------
Table 3.  Mean Spike Recoveries of Nitrate in Three Groundwater Samples


          Analyte Spike             Spike               Rel.  Std.
          Concentration            Recovery             Deviation
             (mg/L)               (percent)             (percent)
               2.00                  113                   10.7
               3.00                  106                    7.6
               5.00                   98                    1.2
              10.0                    89                    2.7
                                   9210 - 7                       Revision 0
                                                                  January 1995

-------
                  METHOD 9210

    POTENTIOMETRIC DETERMINATION OF NITRATE
IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
                i
7.1.1 - 7.1.2
Calibrate Nitrate
ISE.
1
r
7.2 Allow standards
to equilibrate to
room temperature.
^
r
7.3 Rinse
eletrodes.
>
r
7.4 Measure
concentration using
electrode meter
and calculate
concentration.
^
r
7.5 Drain
reference electrode
and clean.
	 3
r
                    9210 - 8
Revision 0
January 1995

-------
                                  METHOD 9211

                    POTENTIOMETRIC DETERMINATION OF BROMIDE
                IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
1.0   SCOPE AND APPLICATION

      1.1   This method can be used for measuring total solubilized bromide in
drinking waters, natural surface waters, groundwaters, domestic and industrial
wastewaters, and in soil extracts (ASTM methods D4646-87, D5233-92 or D3987-85).

            NOTE:  This method is for the  analysis of  simple bromide ion rather
            than total bromide,  as analysis using the ion-selective  electrode is
            not preceded by a distillation step.

      1.2   The method detection limit is 0.2 mg/L.   Bromide concentrations from
0.1 to 1,000 mg/L  may be measured.   However,  when  using a linear calibration,
results less than  0.2 mg/L may be biased up to approximately 40 percent high.

      1.3   ISEs  must  be  used  carefully,  and  results  must  be  interpreted
cautiously, since  an  ISE may  be affected  by numerous analytical interferences
which may  either  increase  or decrease the  apparent analyte  concentration,  or
which may  damage  the  ISE.   Effects of most interferences  can be  minimized or
eliminated by adding appropriate chemical reagents to the sample.  Obtaining the
most  accurate  results,  therefore,   requires  some  knowledge  of  the  sample
composition.

            NOTE:  ISE manufacturers usually include a list of interferences in
            the instruction manual accompanying an  ISE, along with recommended
            methods   for  minimizing  or   eliminating   effects   of   these
            interferences.

2.0   SUMMARY OF METHOD

      2.1   Total  solubilized bromide is determined potentiometrically using a
bromide ion-selective electrode (ISE)  in  conjunction with  a double-junction
reference electrode and a pH  meter  with an  expanded millivolt scale  or an ISE
meter capable of being calibrated directly in terms of bromide concentration.

      2.2   Standards   and   samples  are  mixed  50:1  with  an  ionic  strength
adjustment solution (ISA).  Calibration is  performed  by analyzing  a  series of
standards  and  plotting mV vs.  bromide  concentration  on  semilog   paper  or  by
calibrating the ion meter directly in terms of bromide concentration.

3.0   INTERFERENCES

      3.1   Some polyvalent  cations  (e.g., Fe*3 and Al+3) at high concentrations
(> 300 mg/L)  interfere by forming complexes with bromide which are not measured
by the bromide ISE.  However, in dilute aqueous  solutions, aluminum and iron(III)
do not form complexes  with bromide ions  and dilute  sulfuric acid has  no effect
on the  bromide  concentration because silver sulfate  is  soluble  and  because
hydrogen bromide is a  strong acid in water.  Chloride  is not an interference at


                                   9211  -  1                       Revision 0
                                                                  January 1995

-------
10 times the mass concentration of bromide since the solubility product of silver
chloride is about  400  times  that of silver bromide.   (See  Table  1.)   Bromide
concentration in the presence of very high concentrations of polyvalent cations
can be measured by  treating  the  sample  with  an  equal  volume of EDTA or by the
method of standard  additions.   Solutions containing much  Fe+3 will  be colored
brownish-red at pH values  of  1 or more;  in  such cases,  phosphoric acid or sodium
phosphate monobasic can be added until  the solution is colorless (assuming the
only coloring reagent is Fe+3), at which point the iron interference will  have
been removed and bromide can be accurately determined by the ISE.

      3.2   Sulfide, cyanide, and  ammonia  interfere  with  the determination by
reacting directly with the ISE.  These interferences can be  removed by acidifying
the sample to a pH of 4 with dilute sulfuric acid.

      3.3   Temperature changes  affect electrode potentials.   Using  an  ISE
calibrated at 22°C, a 20.0 mg/L chloride solution was measured as 20.2 mg/L at
22°C and 14.2 mg/L at 32°  (see Ref. 4).  Therefore, standards and samples must
be equilibrated at the same temperature (+ 1°C).

      3.4   The user should  be aware of the  potential  of intereferences  from
colloidal substances and that, if necessary, the samples may be filtered.

4.0   APPARATUS AND MATERIALS

      4.1   pH/mV meter capable of reading to 0.1 mV or an ISE meter.

      4.2   Bromide ISE (Orion 9435 or equivalent) and double-junction reference
electrode (Orion 9002 or equivalent).

      4.3   Thermally isolated magnetic stirrer, Teflon®-coated  stir bar,  and
stopwatch.

      4.4   Volumetric flask, 100 mL.

5.0   REAGENTS

      5.1   Reagent grade  chemicals shall be used in all tests. Unless otherwise
indicated, it is intended  that all  reagents  shall  conform to the specifications
of the Committee on Analytical Reagents of  the American Chemical Society, where
such specifications  are available.   Other grades may  be  used,  provided it is
first ascertained that the reagent is of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent  water.   All references  to  water in this  method  refer to
reagent water, as defined in Chapter One.

      5.3   ISA  solution  (5M NaN03):   Dissolve  42.5  g  of  sodium  nitrate in
sufficient reagent water to make 100 mL of  solution.   Store  in a clean glass or
plastic container.

      5.4   Ethylenediaminetetraacetate (EDTA),  disodium salt (2% C10H12N208Na2).
Use as directed in  Step 3.1.
                                   9211 - 2                       Revision 0
                                                                  January 1995

-------
      5.5   Phosphoric acid (H3P04).   Use  as  directed  in  Step  3.1.

      5.6   Sodium phosphate monobasic (NaH2P04 • H20).  Use as directed in Step
3.1.

      5.7   Dilute sulfuric acid (0.01 M,  H2S04):  Use as directed in Step 3.2.

      5.8   Bromide calibration stock solution (1,000 mg/L Br"): Dissolve 0.1489
g of potassium bromide (dried two hours  at 110'C  and stored in a desiccator) in
reagent water  and  dilute to 100 ml  in  a  volumetric flask.   Store  in  a clean
bottle.

      5.9   Bromide  standard  solution  (100  mg/L Br"):    Dilute  10.0 ml  of
1,000 mg/L bromide calibration stock solution to 100 mL with reagent water in a
volumetric flask.

      5.10  Bromide  calibration  standards:   Prepare  a  series  of calibration
standards by diluting the 100 mg/L bromide standard.   A suitable series  is given
in the table below.

            mL of 100 mg/L          Concentration when Diluted
              Br" Solution              to 50.0 mL  (mg/L Br")
                0.150                           0.300
                0.500                           1.00
                1.50                            3.00
                5.00                           10.0
               15.0                            30.0
6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must have been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   Samples should be stored at 4° C prior  to analysis.

7.0   PROCEDURE

      7.1   Calibration

            7.1.1 When  using a  bromide  ISE and  a  separate  double-junction
      reference  electrode,  ensure that  reference  electrode  inner  and outer
      chambers are filled with the solutions recommended by the manufacturer.
      Equilibrate the electrodes for  at  least one  hour in  a 3.00 mg/L bromide
      standard before use.

            7.1.2 Calibrate  the  bromide  ISE  using  standards   that  narrowly
      bracket the expected sample concentration.   If the sample concentration is
      unknown, calibrate with 1.00 mg/L  and  10.0 mg/L bromide standards.  Add
      50.0 mL of standard and 1.00 mL of ISA to a 100-mL beaker.   Add  a Teflon®-
      coated magnetic stir bar,  place the beaker on a magnetic stir plate,  and


                                   9211 - 3                       Revision 0
                                                                  January 1995

-------
      stir at slow  speed  (no visible vortex).  Immerse  the  electrode  tips to
      just above the rotating stir bar.   If  using  an  ISE meter,  calibrate the
      meter  in  terms  of  bromide concentration  following the  manufacturer's
      instructions.   If using a  pH/mV meter,  record the  meter reading  (mV) as
      soon as the reading  is stable,  but  in no case should the time exceed five
      minutes after immersing the electrode tips.   Prepare a calibration curve
      by plotting measured  potential  (mV) as a  function of  the  logarithm of
      bromide concentration.  The slope must be  54-60  mV per decade of bromide
      concentration.  If the slope is not acceptable, the  ISE  may not be working
      properly.   For corrective action,  consult  the ISE operating manual.

      7.2   Allow samples and standards to equilibrate to room temperature.

      7.3   Prior to and between  analyses, rinse the electrodes thoroughly with
reagent water and gently  shake  off excess water.  Low-level  measurements are
faster if  the  electrode tips are first  immersed  for  five minutes  in  reagent
water.

      7.4   Add 50.0 ml of sample and 1.00 ml of ISA to a  100 mL beaker.  Add a
Teflon®-coated magnetic stir bar.  Place the beaker  on  a magnetic stir plate and
stir at a  slow speed (no  visible vortex).   Immerse the  electrode tips  to just
above the rotating stir bar.  Record  the  meter reading (mV or concentration) as
soon as the reading  is stable, but in  no case should  the time exceed five minutes
after  immersing  the  electrode   tips.    If reading  mV,  determine  bromide
concentration from the calibration curve.

      7.5   When analyses have been completed, rinse  the electrodes thoroughly
and store them  in a  3.00 mg/L bromide standard solution.   If the electrodes will
not be used more  than  one day, drain  the reference electrode internal  filling
solutions, rinse with reagent water,  and store dry.

8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality  control procedures.

      8.2   Initial  Calibration Verification standard  (ICV):   After performing
the calibration  step  (7.1), verify  calibration  by  analyzing  an  ICV.   The ICV
contains  a known bromide  concentration  at  the  mid-range of  the calibration
standards  and  is from  an  independent  source.   ICV  recovery must  be 90-110
percent.   If not, the error  source must be found and  corrected.  An acceptable
ICV must  be analyzed  prior to  sample  analysis.   The  ICV  also serves  as  a
laboratory control  sample.

      8.3   Continuing Calibration Verification  standard  (CCV):  After every 10
samples, and after  the final sample,  a CCV must  be  analyzed.   The CCV contains
a known bromide concentration at mid-calibration  range.  CCV recovery must  be 90-
110 percent.   If not,  the  error  source  must be found and corrected.   If ISE
calibration has changed, all samples analyzed since  the last acceptable CCV must
be re-analyzed.

      8.4   Reagent blank:   After the ICV and after every CCV, a reagent blank
must be analyzed.   A reagent blank  is  reagent water mixed 50:1  with ISA.  The
indicated  reagent blank concentration must be less than  0.3  mg/L bromide.  If


                                   9211  - 4                       Revision 0
                                                                  January  1995

-------
not, the contamination source must be found and corrected.  All samples analyzed
since the last acceptable reagent blank must be re-analyzed.

      8.5   Matrix spike:  Follow the matrix spike protocols presented in Chapter
One.  The spike concentration must be 10 times the detection limit and the volume
added must be  negligible (less than or equal to one-thousandth the sample aliquot
volume).   Spike recovery must be  75-125  percent.    If not, samples  must be
analyzed by the method of standard additions.

9.0   METHOD  PERFORMANCE

      9.1   In a single-laboratory evaluation,  a series  of standards  with known
bromide  concentrations  was   analyzed with  a bromide  ISE.   Measurements  were
conducted over three  consecutive  days  using an Orion 9435  bromide ISE and an
Orion 9002  double-junction  reference  electrode connected to  an Orion  940 ISE
meter.  A two-point calibration (4.00 and 40.0 mg/L bromide)  was  performed prior
to analysis.  Results are listed in Table 2.

      9.2   In the same study,  four groundwater samples were  spiked  with bromide
at four different concentrations and were measured with the bromide  ISE.   (The
groundwater samples  initially contained  <0.1-0.2 mg/L  bromide.)   Each spiked
sample was analyzed at each  concentration,  and  the mean  recoveries  and RSDs are
given in Table 3.

      9.3   A 50 g portion of soil, which initially contained <0.1 mg/kg bromide,
was spiked with 5.00 mg/kg bromide to obtain  an  anion concentration  in a single
extract volume within  the linear range  of the  ISE.  The extract was then analyzed
for bromide using this ISE method, and 92% of the soil spike was recovered.

10.0  REFERENCES

      1.    Franson, Mary Ann  H.,  Ed.   Standard Methods for the Examination of
Water  and  Wastewater,  18th  Edition.    American  Public  Health  Association,
Washington, DC, 1992.

      2.    Model 94-35 Bromide Electrode Instruction Manual.  Orion Research,
Inc., Boston,  MA, 1986.

      3.    Miller, E.L., Waltman,  D.W., and Hillman,  D.C.   Single-Laboratory
Evaluation of  Fluoride, Chloride,  Bromide,  Cyanide,  and Nitrate Ion-Selective
Electrodes for Use  in  SW-846 Methods.  Lockheed  Engineering and Sciences Company
for  Environmental  Monitoring  Systems  Laboratory, U.S.  EPA.   September 1990.
EPA/600/X-90/221.

      4.    Cotton,  F.  Albert,  and  Geoffrey  Wilkinson;  Advanced  Inorganic
Chemistry, 2nd Edition; Interscience Publishers, New York, NY; 1966.

      5.    Weast,  Robert C.,  Ed.; CRC Handbook of Chemistry and Physics,  58th
Edition; CRC Press, Inc.,  Cleveland, Ohio;  1977.

      6.    Kolthoff, I.M.,  E.B. Sandell, E.J. Meehan,  and  Stanley Bruckenstein;
Quantitative Chemical  Analysis, 4th Edition; The MacMillan Company, New York, NY;
1969.
                                   9211 - 5                       Revision 0
                                                                  January 1995

-------
              Table 1.  Bromide ISE Interferences
Analyte
Concentration
(mg/L)
20.0
200
20.0
200
20.0
200
200
20.0
Interference
None
None
300 mg/L Fe+3
300 mg/L Fe+3
300 mg/L Al+3
300 mg/L Al+3
0.01 M H2S04
200 mg/L CV
Measured
Concentration
(mg/L)
20.0
200
19.6
196
20.0
196
201
20.1
RSD
(%)
1.3
1.3
2.0
0.8
3.2
2.0
1.5
1.9
Table 2.  Results from a Single-Laboratory Accuracy Evaluation
                       of a Bromide ISE.
Bromide
Concentration
(mg/L)
0.100
0.200
0.400
1.00
2.00
4.00
10.0
20.0
40.0
100.
200.
400.
1,000.
Bromide
Detected
(mg/L)
0.141
0.217
0.40
1.00
2.02
4.0
10.0
20.3
40.
104.
203.
400.
990.
Bromide
Recovery
(percent)
141
109
100
100
101
100
100
102
100
104
102
100
99
Rel. Std.
Deviation
(percent)
17
6
3
1
2
3
3
8
2
4
6
5
6
                            9211  -  6
Revision 0
January 1995

-------
Table 3.  Mean Spike Recoveries of Bromide in 4 Groundwater Samples


     Analyte  Spike              Spike                Rel.  Std.
     Concentration             Recovery              Deviation
        (mg/L)                (percent)              (percent)
          0.500                   96                    12.2
          1.00                    94                     1.2
          3.00                   101                     2.3
          5.00                    96                     1.0
                             9211  - 7                       Revision 0
                                                            January 1995

-------
                  METHOD 9211

    POTENTIOMETRIC DETERMINATION OF BROMIDE
IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
^
r
7.1.1 - 7.1.2
Calibrate Bromide
ISE.
^
r
7.2 Allow standards
to equilibrate to
room temperature.
^
r
7.3 Rinse
eletrodes.
>
r
7.4 Measure
concentration using
electrode meter
and calculate
concentration.
•^
r
7.5 Drain
reference electrode
and clean.
i
r
                    9211 - 8
Revision 0
January 1995

-------
                                  METHOD 9212

                   POTENTIOMETRIC DETERMINATION OF CHLORIDE
                IN AQUEOUS SAMPLES WITH ION-SELECTIVE  ELECTRODE
1.0   SCOPE AND APPLICATION

      1.1   This method can be used for measuring total solubilized chloride in
drinking waters, natural surface waters, groundwaters, domestic and industrial
wastewaters, and in soil extracts (ASTM methods D4646-87,  D5233-92 or D3987-85).

            NOTE:  This method is for the analysis of simple chloride ion rather
            than total chloride, as analysis using the ion-selective electrode
            is not preceded by a distillation step.

      1.2   The method detection limit is 2.0 mg/L.  Chloride concentrations from
0.4 to 1,000 mg/L  may be measured.   However,  when  using  a linear calibration,
results less than 2.0 mg/L may be biased up to approximately 60 percent high.

      1.3   ISEs  must  be  used  carefully  and  results   must  be  interpreted
cautiously, since  an  ISE may  be  affected by numerous analytical  interferences
which may  either  increase  or decrease the  apparent  analyte  concentration,  or
which may  damage  the  ISE.   Effects of most interferences can  be  minimized  or
eliminated by adding appropriate chemical reagents to the  sample.  Obtaining the
most  accurate  results,  therefore,  requires  some  knowledge  of  the  sample
composition.

            NOTE:  ISE manufacturers usually include  a list of interferences in
            the instruction manual  accompanying an ISE,  along with recommended
            methods   for   minimizing   or   eliminating   effects   of   these
            interferences.

      1.4   The chloride ISE  should not be used in  solutions  containing high
concentrations of bromide or  iodide, cyanide,  or sulfide since these ions form
silver salts less  soluble than silver chloride.  The chloride  ISE will also give
erroneous  readings and will be damaged when used  in solutions  containing free
ammonia,  since such solutions dissolve silver chloride.

2.0   SUMMARY OF METHOD

      2.1   Total solubilized chloride  is determined  potentiometrically using a
chloride ion-selective  electrode  (ISE)  in conjunction with  a  double-junction
reference  electrode,  or a chloride  combination  ISE, and a  pH meter  with  an
expanded millivolt scale or an ISE meter capable of  being  calibrated directly in
terms of chloride concentration.

      2.2   Standards  and  samples  are  mixed  50:1  with an  ionic  strength
adjustment solution (ISA).  Calibration  is  performed by  analyzing  a series  of
standards  and  plotting mV  vs. chloride  concentration  on semi log paper  or  by
calibrating the ion meter directly in terms of chloride concentration.
                                   9212 - 1                       Revision 0
                                                                  January 1995

-------
3.0   INTERFERENCES

      3.1   Polyvalent  cations  (e.g.,  Fe+3  and  Al+3)  interfere  by  forming
complexes with chloride which are not measured by the chloride ISE.  (See Table
1.)  Aluminum and  iron(III)  ions  (at concentrations  of 300 mg/L and higher) form
complexes with chloride  ions  having  approximately  equal  stability.   (A sample
with high solubilized  Fe+3  and Al    concentrations  will  be distinctly colored
yellow-brown.)  Dilute sulfuric acid has no effect on the  chloride concentration
because silver sulfate is soluble, and because  hydrogen chloride is a  strong acid
(completely ionized) in water.  Chloride concentration  in the presence of high
concentrations of polyvalent cations can be measured by  treating the  sample with
an equal volume of EDTA or by the method of standard additions.

      3.2   Bromide,   sulfide,   cyanide,   and  ammonia  interfere   with  the
determination by reacting directly with and damaging the ISE.  Sulfide, cyanide,
and ammonia can be  removed  by acidifying the sample to  a  pH  of  4 with dilute
sulfuric acid.   Bromide and  iodide  can  be removed by  treating  the acidified
sample with potassium bromate, which converts the ions  to bromine and iodate.

      3.3   Temperature  changes  affect  electrode  potentials.   Using  an  ISE
calibrated at 22°C, a 40.0 mg/L chloride solution was measured as 40.0 mg/L at
22°C and 24.8 mg/L at 32° (see Ref. 4).  Therefore, standards and samples must
be equilibrated at the same temperature (+ 1°C).

      3.4   The user  should be aware of the  potential  of  intereferences from
colloidal substances and that, if necessary, the samples may be filtered.

4.0   APPARATUS AND MATERIALS

      4.1   pH/mV meter capable of reading to 0.1 mV or  an ISE meter.

      4.2   Combination  chloride  ISE (Orion 9617 or equivalent),  or separate
chloride ISE (Orion 9417 or  equivalent) and double-junction reference electrode
(Orion 9002 or equivalent).

      4.3   Thermally  isolated magnetic  stirrer, Teflon®-coated  stir bar,  and
stopwatch.

      4.4   Volumetric flask,  100 mL.

5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in  all  tests.  Unless otherwise
indicated, it is intended that all  reagents  shall conform to the specifications
of the Committee on Analytical Reagents of  the American Chemical Society, where
such specifications  are  available.   Other grades may  be used,  provided it is
first ascertained that the reagent is of  sufficiently high  purity  to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent  water.   All  references to  water in this method  refer to
reagent water, as defined in Chapter One.
                                   9212 - 2                       Revision 0
                                                                  January 1995

-------
      5.3   ISA  solution  (5M NaN03):   Dissolve 42.5 g  of sodium  nitrate in
sufficient reagent water to make 100 ml of  solution.  Store in a clean glass or
plastic container.

      5.4   Ethylenediaminetetraacetate (EDTA),  disodium salt (2% C10H12N208Na2).
Use as directed in Step 3.1.

      5.5   Dilute sulfuric acid (0.01 M H2S04):   use as directed  in Step  3.2.

      5.6   Potassium bromate (1%, KBr03):   use as directed in Step 3.2.

      5.7   Chloride  calibration  stock  solution  (1,000 mg/L  CV):   Dissolve
0.1649 g of sodium chloride (dried two hours at HO'C and stored in a desiccator)
in reagent water and dilute to 100 ml in a volumetric flask.   Store in a clean
bottle.

      5.8   Chloride calibration  standards:   Prepare a series  of calibration
standards by diluting the  1,000 mg/L chloride  standard.   A suitable series is
given in the table below.

            ml of 1,000 mg/L          Concentration when Diluted
              CV Solution               to 50.0 mL  (mg/L CV)
               0.050                               1.00
               0.150                               3.00
               0.50                               10.0
               1.50                               30.0
               5.0                               100.0

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All samples  must have been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   Samples should be stored at 4°C.  There are no other special sample
            handling or preservation requirements.

7.0   PROCEDURE

      7.1   Calibration

            7.1.1  If using a chloride  combination  ISE,  ensure  that the ISE is
      filled with  the  solution recommended by  the manufacturer.   Change  the
      solution if  the ISE has not been used for a week.   If using a chloride ISE
      and a separate double-junction  reference electrode, ensure that reference
      electrode inner and outer chambers are filled with solutions recommended
      by the manufacturer.  In  either case, equilibrate the electrode(s) for at
      least one hour in a 30.0 mg/L chloride standard before use.

            7.1.2  Calibrate  the  chloride  ISE  using  standards that  narrowly
      bracket the  expected sample concentration.  If the  sample concentration is


                                   9212 - 3                       Revision 0
                                                                  January 1995

-------
      unknown, calibrate with 10.0 mg/L  and  100  mg/L  chloride  standards.   Add
      50.0 mL of standard and  1.00 mL of  ISA to a 100 ml beaker.  Add a Teflon®-
      coated magnetic stir bar,  place the beaker on a  magnetic stir plate, and
      stir at slow  speed (no  visible vortex).   Immerse  the  electrode  tips to
      just above the rotating stir bar.   If  using  an  ISE meter,  calibrate the
      meter  in  terms of  chloride  concentration following the  manufacturer's
      instructions.  If  using a pH/mV meter,  record the meter  reading  (mV) as
      soon as the reading is stable,  but  in no case  should the time exceed five
      minutes after immersing the electrode tips.  Prepare a calibration curve
      by plotting  measured  potential  (mV) as  a  function of the  logarithm of
      chloride concentration.  The slope must be 54-60  mV per decade of chloride
      concentration.  If the slope is not acceptable, the ISE may not be working
      properly.   For corrective action,  consult the ISE operating manual.

      7.2   Allow samples and standards to equilibrate to room temperature.

      7.3   Prior to and between analyses, rinse  the electrodes thoroughly with
reagent water and  gently shake  off excess water.   Low-level  measurements are
faster if  the electrode tips are first  immersed for   five minutes  in  reagent
water.

      7.4   Add 50.0 mL of sample  and 1.00 mL of  ISA to a  100 mL beaker.  Add a
Teflon®-coated magnetic  stir bar.  Place the  beaker  on  a magnetic stir plate and
stir at a slow speed (no visible vortex).  Immerse the  electrode tip(s)  to just
above the rotating stir bar.  Record  the  meter reading  (mV or concentration) as
soon as the reading  is stable,  but  in no case should the time exceed five minutes
after  immersing  the electrode  tips.    If  reading   mV,  determine  chloride
concentration from the calibration curve.

      7.5   When analyses have been completed, rinse the electrodes thoroughly
and store them in a 30.0 mg/L chloride standard solution.  If the electrodes will
not be used more than one day, drain  the  internal filling  solutions, rinse with
reagent water, and store dry.

8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality  control procedures.

      8.2   Initial Calibration Verification standard   (ICV):   After performing
the calibration step  (7.1),  verify  calibration by  analyzing an  ICV.   The ICV
contains a  known  chloride  concentration  at  the mid-range of  the calibration
standards  and is from  an  independent source.   ICV  recovery  must  be  90-110
percent.  If not, the error source must be found and corrected.  An acceptable
ICV must  be  analyzed  prior to  sample  analysis.   The  ICV  also  serves  as  a
laboratory control sample.

      8.3   Continuing Calibration Verification standard (CCV):  After every 10
samples, and after the final sample,  a CCV must be analyzed.   The CCV contains
a known chloride concentration at mid-calibration range.  CCV recovery must be
90-110 percent.   If not, the error source must be found and corrected.   If ISE
calibration has  changed,  all samples analyzed since the last acceptable CCV must
be re-analyzed.


                                   9212  - 4                       Revision 0
                                                                  January  1995

-------
      8.4   Reagent blank:  After the ICV and after every CCV, a reagent blank
must be analyzed.  A  reagent  blank  is  reagent  water mixed 50:1 with ISA.  The
indicated reagent blank concentration must be less than 1 mg/L chloride.   If not,
the contamination source must be found and corrected.  All  samples  analyzed since
the last acceptable reagent blank must be re-analyzed.

      8.5   Matrix spike: Follow the matrix  spike protocols presented in Chapter
One.  The spike concentration must be 10 times the detection limit  and the volume
added must  be negligible (less than or equal to one-thousandth the  sample aliquot
volume).   Spike  recovery  must be  75-125  percent.    If  not,  samples  must be
analyzed by the method  of standard  additions.

9.0   METHOD PERFORMANCE

      9.1   In a single-laboratory evaluation,  a series of standards with known
chloride concentrations was  analyzed with a chloride  ISE.   Measurements were
conducted over three consecutive days using an Orion 9609 chloride  combination
ISE connected to an Orion 940  ISE meter.  A two-point calibration  (4.CO and 40.0
mg/L chloride)  was performed prior to analysis.  The  results are listed in Table
2.

      9.2   In the same study,  six groundwater samples were spiked with chloride
at four different concentrations and were measured with the  chloride ISE.  (The
groundwater samples  initially contained 1.3-23 mg/L chloride.)    Each  spiked
sample was analyzed at  each concentration and the mean recoveries and RSDs are
given in Table 3.

      9.3   A  50  g  portion   of  soil, which  initially  contained  17.3  mg/kg
chloride,  was spiked with 25.0 mg/kg chloride to obtain an  anion concentration
in a single extract volume within the linear range of the  ISE.  The  extract was
then analyzed for chloride using  this ISE method, and 109% of the  soil spike was
recovered.

10.0  REFERENCES

      1.    Franson,  Mary Ann H., Ed.  Standard Methods for the  Examination of
Water  and  Wastewater,   18th  Edition.    American  Public  Health Association,
Washington, DC, 1992.

      2.    Model 96-17 Chloride Combination Electrode Instruction Manual.  Orion
Research,  Inc., Boston, MA, 1988.

      3.    Miller, E.L., Waltman,  D.W., and  Hillman,  D.C.   Single-Laboratory
Evaluation of Fluoride,  Chloride,  Bromide,  Cyanide,  and  Nitrate  Ion-Selective
Electrodes for  Use in SW-846 Methods.  Lockheed  Engineering and  Sciences Company
for Environmental  Monitoring  Systems Laboratory, U.S.  EPA.   September 1990.
EPA/600/X-90/221.

      4.    Cotton,  F.   Albert,  and  Geoffrey  Wilkinson;   Advanced Inorganic
Chemistry,  2nd Edition;  Interscience Publishers, New York, NY; 1966.

      5.    Weast  Robert C.,   Ed.; CRC Handbook  of Chemistry and  Physics, 58th
Edition; CRC Press, Inc., Cleveland, Ohio;  1977.

                                   9212 -  5                       Revision 0
                                                                  January 1995

-------
      6.    Kolthoff, I.M.,  E.B. Sandell, E.J. Meehan,  and Stanley Bruckenstein;
Quantitative Chemical Analysis, 4th Edition; The MacMillan Company,  New York, NY;
1969.
                                                                                    i
                                   9212  - 6                       Revision 0
                                                                  January 1995

-------
             Table 1.  Chloride ISE Interferences
Analyte
Concentration
(mg/L)
20.0
200.0
20.0
200.0
20.0
200.0
200.0
200.0
20.0
Interference
None
None
300 mg/L Fe+3
300 mg/L Fe+3
300 mg/L Al+3
300 mg/L Al+3
0.01 M H2S04
200 mg/L CN' +
0.01 M H2S04
200 mg/L S'2 +
0.01 M H2S04
Measured
Concentration
(mg/L)
19.9
200.0
19.6
183.0
19.3
175.0
201.0
198.0
19.9
RSD
(%)
2.2
1.3
3.0
4.5
3.3
6.2
1.5
1.0
0.5
Table 2.  Results from a single-laboratory accuracy evaluation
                       of  a  chloride  ISE
Chloride
Concentration
(mg/L)
0.400
1.00
2.00
4.00
10.0
20.0
40.0
100
200
400
1,000
Chloride
Detected
(mg/L)
0.64
1.32
2.07
4.0
10.0
19.4
40.0
100
201
390
970
Chloride
Recovery
(percent)
160
132
104
100
100
97
100
100
101
99
97
Rel. Std.
Deviation
(percent)
21
9
4
3
4
7
3
4
1
4
3
                           9212 - 7
Revision 0
January 1995

-------
Table 3.  Mean Spike Recoveries of Chloride in 6 Groundwater Samples
     Analyte Spike             Spike               Rel.  Std.
     Concentration            Recovery             Deviation
        (mg/L)               (percent)             (percent)
          2.00                  107                    8.2
          3.00                   95                    3.6
          5.00                   93                    3.6
         10.0                   102                    4.3
i
                              9212 - 8                       Revision 0
                                                             January 1995

-------
                  METHOD 9212
   POTENTIQMETRIC DETERMINATION OF CHLORIDE
IN AQUEOUS SAMPLES WITH ION-SELECTIVE  ELECTRODE
>
r
7.1.1 - 7.1.2
Calibrate Chloride
ISE.
^
r
7.2 Allow standards
to equilibrate to
room temperature.
^
r
7.3 Rinse
eletrodes.
i
r
7.4 Measure
concentration using
electrode meter
and calculate
concentration.
i
r
7.5 Drain
reference electrode
and clean.
^
r
                   9212 - 9
Revision 0
January 1995

-------
                                  METHOD 9213

                    POTENTIOHETRIC DETERMINATION OF CYANIDE
        IN AQUEOUS SAMPLES AND DISTILLATES WITH ION-SELECTIVE ELECTRODE
1.0   SCOPE AND APPLICATION

      1.1   This method can be used for measuring free (non-complexed) cyanide
and hydrocyanic acid  in  drinking  waters,  natural  surface waters, domestic and
industrial wastewaters, and in soil extracts (ASTM methods D4646-87, D5233-92 or
D3987-85).   The  method may also  be  used  as the  determinative  step  for total
cyanide in distillate following the distillation  in Method 9010.

      1.2   The method detection limit is 0.05 mg/L.  Cyanide concentrations from
0.01 to 10 mg/L may be measured.  However, using  a linear calibration, results
less than 0.05 mg/L may be biased up to approximately 120 percent high.

      1.3   ISEs  must be  used  carefully,  and   results  must  be  interpreted
cautiously, since an  ISE may  be  affected  by numerous analytical interferences
which may  either  increase  or decrease the  apparent  analyte  concentration,  or
which may  damage  the  ISE.   Effects of most interferences  can be minimized or
eliminated by adding appropriate chemical reagents to the  sample. Obtaining the
most  accurate  results,  therefore,  requires  some  knowledge  of  the  sample
composition.

            NOTE: ISE manufacturers usually include a list of interferences in
            the instruction manual accompanying an ISE, along with recommended
            methods   for   minimizing   or   eliminating    effects   of   these
            interferences.

      1.4   The cyanide ISE should not be used in  solutions  containing more than
30 mg/L of free  cyanide,  because such solutions  will  dissolve  the  ISE sensor
silver compounds.

2.0   SUMMARY OF METHOD

      2.1   Cyanide  is  determined  potentiometrically  using  a cyanide  ion-
selective  electrode  (ISE)  in conjunction  with  a double-junction  reference
electrode and a pH meter with an expanded millivolt scale or an ISE meter capable
of being calibrated directly  in terms of cyanide  concentration.

      2.2   Standards  and   samples  are mixed  100:1  with  an  ionic  strength
adjustment solution (ISA).  Calibration is  performed  by  analyzing  a  series of
standards  and  plotting mV  vs.  cyanide concentration  on  semilog paper  or  by
calibrating the ion meter directly in terms of cyanide concentration.

3.0   INTERFERENCES

      3.1   Transition metal cations  interfere  by  forming very stable complexes
with cyanide which are not measured by the  cyanide ISE.  For example,  copper(II)
ions interfere with the cyanide determination by oxidizing cyanide to  cyanogen
gas and precipitating  copper(I) cyanide. Because these complexes are very stable
and their rate of dissociation is slow (days)  at room temperature (25°C), this

                                   9213 -  1                       Revision 0
                                                                  January 1995

-------
method is restricted  to  the analysis of free cyanide  and  hydrocyanic acid in
undistilled samples.   Total cyanide may be determined in samples distilled as per
Method 9010.  (Distillation removes interferants and breaks down metal-analyte
complexes and insoluble salts.)

      3.2   Several anions  and one  acid were  tested  for cyanide interference.
(See Table 1.)  As shown,  hydrogen  ions  (acid) interfere by forming hydrocyanic
acid.  Hydrogen ions  are removed by adding ISA.

      3.3   Sulfide interferes with the determination by reacting directly with
the ISE.   This interference can be removed after adding ISA by adding and well
mixing small amounts  (about 0.05 g) of powdered lead carbonate until the added
powder remains white.

      3.4   Temperature  changes  affect electrode  potentials.   Using  an  ISE
calibrated at 22 C, a 1.00 mg/L cyanide solution was measured  as  0.98 mg/L at 22
°C and 0.64  mg/L at 32°C  (see Ref. 4).  Therefore, standards and  samples must be
equilibrated at the same temperature  ( + 1°C).

      3.5   The user  should be aware of the  potential  of  intereferences from
colloidal substances  and that, if necessary, the samples may be filtered.

4.0   APPARATUS AND MATERIALS

      4.1   pH/mV meter capable of reading to 0.1 mV or an ISE meter.

      4.2   Cyanide ISE (Orion  9406 or equivalent) and double-junction reference
electrode (Orion 9002 or equivalent).

      4.3   Thermally  isolated magnetic stirrer, Teflon®-coated stir bar,  and
stopwatch.

      4.4   Volumetric flask, 100 mL.

5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such  specifications  are  available.  Other grades may  be used,  provided it is
first ascertained that the reagent is of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent  water.   All  references to  water  in this method  refer to
reagent water, as defined in Chapter One.

      5.3   ISA  solution  (10M NaOH):   Dissolve 40 g  of sodium  hydroxide in
sufficient  reagent water to make  100 mL  of  solution.   Cool,  and  store  in  a
polyethylene bottle.

CAUTION:   This solution is  extremely corrosive.
                                   9213 - 2                       Revision 0
                                                                  January 1995

-------
      5.4   Powdered lead carbonate [(PbC03)2 • Pb(OH)2]:  use in approximately
0.05 g increments to remove sulfide interference.  (See Step 3.3.)

      5.5   Cyanide  calibration stock  solution   (1,000  mg/L CN"):   Dissolve
0.2504 g  of potassium  cyanide  (dried  two hours  at  110°C and  stored  in  a
desiccator) in  reagent  water,  add  1.0 ml  of  ISA and dilute to  100.0  ml in a
volumetric flask.  Store in a clean bottle.

      5.6   Cyanide  standard  solution (100.0 mg/L  CN"):   Dilute  10.00  ml of
1,000 mg/L cyanide calibration stock solution to  100.0 mL with reagent water in
a volumetric flask.

      5.7   Cyanide  calibration  standards:  Prepare a series  of calibration
standards by diluting the  100.0 mg/L cyanide standard.  A  suitable  series is
given in the table below.

            mL of 100.0 mg/L             Concentration when Diluted
              CM' Solution                to 50.0 mL   (mg/L  CN')
                0.0150                           0.0300
                0.0500                           0.100
                0.150                            0.300
                0.500                            1.00
                1.50                             3.00

      5.8   Sodium hydroxide solution (50%), NaOH.   Commercially available.

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All samples  must have been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   Samples must be preserved by adding 50% sodium hydroxide until the
pH is greater than or equal to 12  at the time of collection and  should be chilled
to 4°C.  When properly preserved,  cyanide samples can be stored for up to 14 days
prior to analysis.

7.0   PROCEDURE

      7.1   Calibration

            7.1.1 When  using a  cyanide  ISE  and  a  separate double-junction
      reference electrode,  ensure that  reference   electrode  inner  and  outer
      chambers  are  filled  with   solutions  recommended  by the  manufacturer.
      Equilibrate the electrodes  for at least one hour in a 0.300 mg/L cyanide
      standard before use.

            7.1.2 Calibrate  the   cyanide  ISE  using  standards  that  narrowly
      bracket the expected sample concentration.  If the  sample concentration is
      unknown,  calibrate with 0.100 mg/L and 1.00 mg/L cyanide standards.  Add
      50.0 mL of standard and 0.50 mL of ISA to a 100 mL  beaker.  Add a Teflon®-
      coated magnetic stir bar,  place the beaker on a magnetic stir plate, and


                                   9213 - 3                       Revision 0
                                                                  January 1995

-------
      stir at  slow  speed  (no visible vortex).   Immerse  the  electrode tips to
      just above the rotating stir  bar.   If using  an ISE meter,  calibrate the
      meter  in terms  of  cyanide concentration  following the  manufacturer's
      instructions.  If using a  pH/mV meter,  record  the  meter reading (mV) as
      soon as the reading  is stable,  but  in  no case should the time exceed five
      minutes after immersing the electrode tips.  Prepare a calibration curve
      by plotting  measured  potential (mV)  as a  function of  the  logarithm of
      cyanide concentration.  The slope must be  54-60 mV per decade of cyanide
      concentration.  If the slope is not acceptable,  the  ISE  may not be working
      properly.  For corrective action,  consult  the ISE operating manual.

      7.2   Allow samples and standards to equilibrate to room temperature.

      7.3   Prior to and between  analyses, rinse  the electrodes thoroughly with
reagent water  and  gently  shake  off  excess water.  Low-level  measurements are
faster if the electrode tips are first immersed  five  minutes in reagent water.

      7.4   Add 50.0 ml of sample and 0.50 ml of  ISA  to a 100 ml beaker.  Add a
Teflon®-coated magnetic stir bar.  Place the beaker  on a magnetic stir plate and
stir at a slow speed (no  visible vortex).   Immerse the  electrode  tips to just
above the rotating stir bar.  Record  the  meter reading (mV or concentration) as
soon as the reading  is stable, but in  no case should  the time exceed five minutes
after  immersing  the  electrode   tips.    If reading  mV,  determine  cyanide
concentration from the calibration curve.

      7.5   When analyses have been completed, rinse  the electrodes thoroughly
and store them in a  0.30 mg/L cyanide standard solution.   If the electrodes will
not be used more  than  one day, drain the reference electrode internal filling
solutions, rinse with reagent water, and store dry.

8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality control procedures.

      8.2   Initial Calibration Verification standard (ICV):   After performing
the calibration  step  (7.1), verify  calibration  by  analyzing  an  ICV.   The ICV
contains  a  known cyanide concentration  at  the  mid-range of  the  calibration
standards and  is from  an  independent  source.   ICV  recovery must  be  90-110
percent.  If not, the error source must be found and  corrected.  An acceptable
ICV must  be  analyzed  prior to  sample  analysis.  The  ICV  also  serves  as  a
laboratory control sample.

      8.3   Continuing Calibration Verification  standard  (CCV):  After every 10
samples, and after the final sample, a CCV must be  analyzed.  The CCV contains
a known cyanide concentration at mid-calibration  range.  CCV recovery must  be 90-
110 percent.  If not,  the  error  source must  be found  and  corrected.  If the ISE
calibration has changed, all samples  analyzed since  the last acceptable CCV must
be re-analyzed.

      8.4   Reagent blank:  After the ICV and after every CCV, a reagent blank
must be analyzed.  A reagent blank is reagent water mixed 100:1 with ISA.  The
indicated reagent blank concentration must be less  than 0.03 mg/L cyanide.  If
not, the contamination  source must be found and corrected. All samples analyzed
since the last acceptable reagent blank must be re-analyzed.

                                   9213  - 4                       Revision 0
                                                                  January  1995

-------
      8.5   Matrix spike:  Follow the matrix spike protocols presented in Chapter
One.  The spike concentration must be 10 times the detection limit and the volume
added must be negligible (less than or equal to one-thousandth the sample aliquot
volume).   Spike recovery must be  75-125  percent.    If  not, samples  must be
analyzed by the method of standard additions.

9.0   METHOD PERFORMANCE

      9.1   In a single-laboratory evaluation,  a  series of standards with known
cyanide  concentrations  was  analyzed with  a cyanide  ISE.    Measurements  were
obtained over three consecutive days using an  Orion 9406 cyanide ISE and an Orion
9002 double-junction reference electrode connected  to an Orion  940 ISE meter.
A  two-point  calibration (0.20 and 2.0  mg/L  cyanide) was  performed  prior to
analysis.  The results are listed  in Table 2.

      9.2   In the  same study, three groundwater samples were spiked with cyanide
at four different concentrations and were measured with the cyanide ISE.   (The
groundwater samples initially contained <0.1  mg/L cyanide.)   Each  spiked sample
was analyzed at each concentration  and the mean recoveries and RSDs are given in
Table 3.

      9.3   A 50g portion of soil, which  initially contained  <0.1 mg/kg cyanide,
was spiked with 2.50 mg/kg cyanide to obtain  an anion  concentration  in a single
extract volume within the linear range of  the  ISE.  The extract was  then analyzed
for cyanide using this ISE method, and 92% of the soil spike was recovered.

10.0  REFERENCES

      1.    Franson,  Mary Ann  H.,  Ed.  Standard Methods for the Examination of
Water  and Wastewater,  18th  Edition.    American  Public Health  Association,
Washington, DC, 1992.

      2.    Model 94-06 Cyanide Electrode Instruction Manual.  Orion Research,
Inc., Boston, MA, 1986.

      3.    Miller, E.L., Waltman,  D.W.,  and Hillman, D.C.   Single-Laboratory
Evaluation of Fluoride, Chloride,  Bromide,  Cyanide,  and  Nitrate  Ion-Selective
Electrodes for Use  in SW-846 Methods.  Lockheed Engineering and Sciences Company
for  Environmental  Monitoring   Systems Laboratory,  U.S.  EPA.   September 1990.
EPA/600/X-90/221.

      4.    Cotton, F.  Albert,  and  Geoffrey Wilkinson;   Advanced Inorganic
Chemistry, 2nd Edition; Interscience Publishers,  New York, NY;  1966.

      5.    Kolthoff,  I.M.,  E.B. Sandell, E.J. Meehan,  and Stanley Bruckenstein;
Quantitative Chemical Analysis, 4th Edition; The MacMillan Company, New York, NY;
1969.
                                   9213 - 5                       Revision 0
                                                                  January 1995

-------
        Table  1.   Cyanide  ISE  Interferences
Analyte
Concentration
(mg/L)
0.200
2.00
2.00
0.200
0.200
2.00
Interference
None
None
0.01 M H2S04
200 mg/L CT
200 mg/L Br"
20 mg/L S'2 + 0.05
g PbCo3
Measured
Concentration
(mg/L)
0.201
2.00
<0.05
0.204
0.200
2.02
RSD
(%)
1.0
0.7
*
2.0
2.7
1.6
* Single Measurement
Table 2.  Results From a Single-Laboratory Accuracy
             Evaluation  of  a  Cyanide  ISE
Cyanide
Concentration
(mg/L)
0.0100
0.0200
0.0500
0.100
0.200
0.500
1.00
2.00
5.00
10.0
Cyanide
Detected
(mg/L)
0.0217
0.0340
0.0520
0.103
0.198
0.48
1.03
2.02
5.00
9.9
Cyanide
Recovery
(percent)
217
170
104
103
99
96
103
101
100
99
Rel. Std.
Deviation
(percent)
27
13
8
6
3
6
6
3
6
4
                      9213 - 6
Revision 0
January 1995

-------
Table 3.  Mean Spike Recoveries of Cyanide in 3 Groundwater Samples
    Analyte  Spike              Spike                Rel.  Std.
    Concentration             Recovery              Deviation
        (mg/L)                (percent)              (percent)
          0.0500                 115                    22.3
          0.150                  103                     7.6
          0.300                   98                     8.3
          1.00                   103                     1.1
                             9213  - 7                        Revision 0
                                                             January 1995

-------
                  METHOD 9213
    POTENTIOMETRIC DETERMINATION OF CYANIDE
IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
^
f
7.1.1 - 7.1.2
Calibrate Cyanide
ISE.
•^
r
7.2 Allow standards
to equilibrate to
room temperature.
^
r
7.3 Rinse
eletrodes.
^
r
7.4 Measure
concentration using
electrode meter
and calculate
concentration.
i
r
7.5 Drain
reference electrode
and clean.
\
r
                     9213 - 8
Revision 0
January 1995

-------
                                  METHOD 9214

                   POTENTIOMETRIC DETERMINATION OF FLUORIDE
                IN AQUEOUS SAMPLES WITH  ION-SELECTIVE  ELECTRODE
1.0   SCOPE AND APPLICATION

      1.1   This method can be used for measuring total solubilized fluoride in
drinking waters, natural surface waters, groundwaters, domestic and industrial
wastewaters, and in soil extracts (ASTM methods D4646-87,  D5233-92 or D3987-85).

            NOTE:  This method is for the analysis of simple  fluoride ion rather
            than total fluoride, as analysis using the ion-selective electrode
            is not preceded by a distillation step.

      1.2   The method detection limit is 0.5 mg/L.  Fluoride  concentrations from
0.025 to 500 mg/L may  be measured.  However,  using  a linear calibration, results
less than 0.5 mg/L may be biased up to approximately 160% high.

      1.3   ISEs  must  be   used  carefully  and  results  must  be  interpreted
cautiously, since  an  ISE may  be  affected by numerous analytical  interferences
which may  either increase  or decrease the  apparent  analyte concentration,  or
which may  damage the  ISE.   Effects of most  interferences can  be  minimized or
eliminated by  adding appropriate chemical reagents to the  sample.  Obtaining the
most  accurate  results,  therefore,  requires  some  knowledge  of  the  sample
composition.

            NOTE:  ISE manufacturers usually include a list  of interferences in
            the  instruction manual  accompanying an ISE,  along with recommended
            methods   for   minimizing    or   eliminating   effects   of   these
            interferences.

2.0   SUMMARY OF METHOD

      2.1   Total solubilized fluoride  is determined  potentiometrically using a
fluoride ion-selective electrode (ISE)  in conjunction with  a standard single-
junction reference  electrode,  or a  fluoride combination ISE,  and a pH meter with
an expanded millivolt  scale  or an ISE meter capable  of being  calibrated directly
in terms of fluoride concentration.

      2.2   Standards  and  samples  are mixed  1:1  with a total  ionic  strength
adjustment buffer  (TISAB).   TISAB adjusts ionic strength, buffers pH to 5-5.5,
and contains a chelating agent to break up metal-fluoride complexes.  Calibration
is performed by  analyzing  a series of standards and plotting  mV  vs.  fluoride
concentration  on semilog paper or by calibrating the ion meter directly in terms
of fluoride concentration.

3.0   INTERFERENCES

      3.1   Polyvalent  cations  (e.g.,  Fe+3 and  Al"1"3)   interfere  by  forming
complexes with fluoride which are not measured by the fluoride ISE.  (See Table


                                   9214  - 1                        Revision 0
                                                                  January 1995

-------
1.)  The results in Table 1 are  in agreement with the  fact that aluminum-fluoride
complex  ions  are  approximately  ten  times  more   stable  than  corresponding
iron(III)-fluoride complex ions.  As  the aluminum concentration increases, more
fluoride is consumed to  form the metal-fluoride  complex.   Adding TISAB, which
contains a strong chelating agent, eliminates this  interference  by complexing
polyvalent cations.

      3.2   Sample pH is  critical.   Hydroxide is  a  significant interferant at
concentrations ten  times  the  fluoride concentration.   This  interference is
avoided by adding TISAB which  buffers  the  sample at  a  pH  of  5-5.5.   At low pH
values, fluoride forms bifluoride (HF2~) which is not detected  by the fluoride
ISE.  Again,  adding TISAB prevents this interference by buffering the pH.

      3.3   Temperature  changes  affect electrode  potentials.    Using  an  ISE
calibrated at 22°C,  a 20.0 mg/L fluoride solution was measured as 20.3 mg/L at
22°C and 13.6 mg/L at 32°C (see Ref.  4).  Therefore,  standards and samples must
be equilibrated at the same temperature (+ 1°C).

      3.4   The user  should be aware of the  potential  of intereferences from
colloidal  substances and that,  if necessary, the samples may be filtered.

4.0   APPARATUS AND MATERIALS

      4.1   pH/mV meter capable of reading to 0.1 mV or an ISE meter.

      4.2   Combination  fluoride  ISE (Orion 9609 or equivalent),  or separate
fluoride ISE (Orion 9409 or equivalent) and reference electrode (Orion 9001 or
equivalent) prepared for use as described in owner's manual.

      4.3   Thermally isolated magnetic  stirrer,  Teflon®-coated  stir bar,  and
stopwatch.

      4.4   Volumetric flask,   1,000 mL.

      4.5   Polyethylene labware.

5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that  all reagents  shall conform to the specifications
of the Committee on  Analytical  Reagents of  the American Chemical Society, where
such specifications  are  available.   Other grades may  be  used,  provided it is
first ascertained that the reagent  is of sufficiently high  purity to permit its
use without lessening the  accuracy of the determination.

      5.2   Reagent  water.   All  references  to  water in this  method refer to
reagent water, as defined  in Chapter One.

      5.3   Sodium hydroxide  solution (5M NaOH):   Dissolve  200 g  of NaOH in
sufficient reagent water to make 1  L of solution.   Store  in a tightly sealed
polyethylene bottle.
                                   9214 - 2                       Revision 0
                                                                  January 1995

-------
CAUTION:  This solution is extremely corrosive.

      5.4   Glacial acetic acid (CH3C02H).

      5.5   Sodium chloride (NaCl).

      5.6   CDTAor 1,2-cyclohexanedinitrilo-tetraacetic acid (C6H10[N(CH2C02H)2]2
• H20).

      5.7   TISAB  solution

            5.7.1 To  approximately 500  mL of  reagent  water  add  57.0 ml of
      glacial acetic acid, 58.0 g of sodium chloride,  and 4.00  g  of  CDTA.   Stir
      to  dissolve  and cool to room temperature.   Adjust the  solution pH to
      between  5.0  and  5.5 with  5M  NaOH   (about  150  ml  will  be  required).
      Transfer the solution to a 1,000 ml  volumetric  flask  and dilute to the
      mark with  reagent water.   Transfer  the  solution to a clean polyethylene
      bottle.

            5.7.2 Alternatively, TISAB solution is available commercially (Orion
      940999 or equivalent).

      5.8   Fluoride  calibration  stock  solution  (1,000 mg/L  F"):    Dissolve
0.2210 g  of  sodium fluoride  (NaF,  dried  two hours  at 110°C  and stored in a
desiccator) in reagent water and dilute to  100 mL in a  polyethylene volumetric
flask.  Store in a clean polyethylene bottle.

      5.9   Fluoride standard  solution (100 mg/L  F"):   Dilute 10.0 mL of 1,000
mg/L  fluoride  calibration stock  solution  to  100 mL  with reagent  water in a
polyethylene volumetric flask.

      5.10  Fluoride calibration  standards:  Prepare  a series of calibration
standards by diluting the 100 mg/L fluoride  standard.  A suitable series  is given
in the table below.

            mL of  100 mg/L             Concentration when Diluted
              F"  Solution                to 50.0  mL   (mg/L  F')
               0.0500                          0.100
               0.150                           0.300
               0.500                           1.00
               1.50                            3.00
               5.00                           10.0
              15.0                            30.0

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  must have been  collected  using a  sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.
                                   9214 - 3                       Revision 0
                                                                  January 1995

-------
      6.2   Samples  must  be  stored  and  handled  in  polyethylene  containers.
Samples should be stored at 4°C.

7.0   PROCEDURE

      7.1   Calibration

            7.1.1 If using a  fluoride  combination  ISE,  ensure  that the ISE is
      filled with  the  solution recommended  by  the manufacturer.   Change the
      solution if the ISE  has  not been  used for a week.  If  using  a  fluoride ISE
      and a separate reference electrode, ensure that the reference  electrode is
      filled with the solution recommended by the manufacturer.  In  either case,
      equilibrate the electrodes for 24 hours in a 10.0 mg/L fluoride standard
      before use.

            7.1.2 Calibrate the  fluoride  ISE using  standards  that  narrowly
      bracket the expected sample concentration.  If the  sample concentration is
      unknown, calibrate with 1.00 mg/L and 10.0 mg/L fluoride standards.  Add
      20.0 mL of standard and 20.0 mL of TISAB to a 50 mL polyethylene beaker.
      Add a Teflon®-coated magnetic  stir  bar, place  the beaker  on a magnetic
      stir plate,  and  stir at  slow  speed  (no  visible  vortex).    Immerse the
      electrode  tips to  just  above  the  rotating stir  bar.   If using  an ISE
      meter, calibrate the meter in  terms  of fluoride concentration following
      the manufacturer's  instructions.  If using a pH/mV  meter, record the meter
      reading (mV) as soon as the reading is stable, but in no case should the
      time exceed five minutes after immersing  the  electrode  tips.  Prepare a
      calibration curve by plotting measured potential (mV)  as  a function of the
      logarithm  of  fluoride   concentration.   The slope must  be  54-60  mV per
      decade of fluoride concentration.  If the slope is  not acceptable, the ISE
      may not  be working  properly.    For  corrective action,  consult  the ISE
      operating manual.

            NOTE: Use  only   plasticware  (polyethylene)  when  preparing  and
            analyzing fluoride samples.

      7.2   Allow samples and standards to equilibrate to room temperature.

      7.3   Prior to and between analyses, rinse the electrode thoroughly with
reagent water and  gently  shake  off excess water.   Low-level  measurements are
faster if  the  electrode  tips are  first  immersed for five minutes in reagent
water.

      7.4   Add 20.0 mL of sample and 20.0 mL of TISAB to a 50 mL polyethylene
beaker.  Add a Teflon®-coated  magnetic stir bar.   Place the beaker on a magnetic
stir plate and stir at  a  slow  speed (no visible vortex).  Immerse the electrode
tips to  just  above the rotating  stir  bar.   Record  the meter reading  (mV or
concentration) as soon  as  the  reading is stable,  but  in  no case should the time
exceed five minutes  after immersing the  electrode tips.  If reading mV, determine
fluoride concentration from the calibration curve.

      7.5   When analyses have been completed, rinse the electrodes thoroughly
and store them in a 10.0 mg/L fluoride standard solution.   If the electrodes will


                                   9214 - 4                        Revision 0
                                                                  January 1995

-------
not be used more than one day, drain the internal filling solution, rinse with
reagent water, and store dry.

8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality control procedures.

      8.2   Initial Calibration Verification standard (ICV):  After performing
the calibration  step  (7.1),  verify calibration by analyzing  an  ICV.   The ICV
contains a known fluoride concentration at mid-range of the calibration standards
and is from an independent source.  ICV recovery must be 90-110 percent.   If not,
the error source must be found and corrected.  An acceptable ICV must be analyzed
prior to sample analysis.  The ICV also serves as a laboratory control sample.

      8.3   Continuing Calibration Verification standard (CCV):  After every 10
samples, and after the final sample, a CCV must be analyzed.  The CCV contains
a known fluoride concentration at mid-calibration range.  CCV recovery must be
90-110 percent.  If not, the error source must be found and corrected.  If ISE
calibration has changed,  all  samples analyzed since the last  acceptable CCV must
be re-analyzed.

      8.4   Reagent blank:  After the ICV and after every CCV, a reagent blank
must be analyzed.  A reagent blank is reagent water mixed 1:1 with TISAB.  The
indicated reagent blank concentration must be less than 0.1 mg/L fluoride.  If
not, the contamination  source must be found and corrected.  All samples analyzed
since the last acceptable reagent blank must be re-analyzed.

      8.5   Matrix spike:  Follow the matrix spike protocols presented in Chapter
One.  The  spike concentration must be 10 times the detection limit  and the volume
added must be negligible (less than or equal to one-thousandth the  sample aliquot
volume).   Spike recovery must  be 75-125  percent.    If not, samples  must  be
analyzed by the method of standard additions.

9.0   METHOD PERFORMANCE

      9.1   In a single-laboratory evaluation,  a  series  of standards with known
fluoride concentrations was  analyzed  with a fluoride  ISE.   Measurements were
conducted over three consecutive days using an Orion 9609 fluoride combination
ISE connected to an Orion 940 ISE meter.  A two-point  calibration  (1.00 and 10.0
mg/L fluoride)  was  performed  prior to analysis.  The results  are listed in Table
2.

      9.2   In  the same  study,  12  groundwater  samples  and six  extraction
procedure (EP,  Method 1310)  soil leachate samples were spiked with fluoride at
four different concentrations and  were measured with the fluoride  ISE.   (The
groundwater samples initially contained 0.1-2 mg/L fluoride and the EP leachates
initially contained 0.1-8 mg/L  fluoride.)   Each  spiked  sample was analyzed at
each concentration and the mean recoveries and RSDs are listed in Table 3.
                                   9214 - 5                       Revision 0
                                                                  January 1995

-------
10.0  REFERENCES

      1.    Franson, Mary Ann H., Ed.  Standard Methods for the Examination of
Water  and Wastewater,  18th  Edition.    American  Public  Health  Association,
Washington, DC, 1992.

      2.    Model 96-09 Fluoride Combination Electrode Instruction Manual.  Orion
Research, Inc., Boston, MA, 1988.

      3.    Methods  for  Chemical   Analysis   of   Water   and   Wastes.     U.S.
Environmental Protection Agency, Cincinnati, Ohio, 1983.

      4.    Miller, E.L., Waltman, D.W.,  and  Hillman,  D.C.   Single-Laboratory
Evaluation of  Fluoride, Chloride,  Bromide,  Cyanide,  and  Nitrate Ion-Selective
Electrodes for Use in SW-846 Methods.  Lockheed Engineering and Sciences Company
for Environmental  Monitoring Systems  Laboratory,  U.S.  EPA.   September 1990.
EPA/600/X-90/221.

      5.    Cotton,  F. Albert,  and  Geoffrey  Wilkinson; Advanced  Inorganic
Chemistry, 2nd Edition; Interscience Publishers, New York, NY; 1966.

      6.    Weast, Robert C., Ed.; CRC Handbook of Chemistry and Physics, 58th
Edition; CRC Press, Inc., Cleveland, Ohio; 1977.
                                   9214 - 6                       Revision 0
                                                                  January 1995

-------
        Table 1.  Fluoride ISE Interferences
Analyte
Concentration
(mg/L)
1.00
1.00
1.00
1.00
1.00
1.00
1.00
1.00
Interference
None
1 mg/L Al+3
5 mg/L Al+3
25 mg/L Al+3
50 mg/L Al+3
1 mg/L Fe+3
5 mg/L Fe+3
50 mg/L Fe+3
Measured
Concentration
(mg/L)
0.95
0.97
0.86
0.51
0.41
1.00
1.02
0.92
RSD
t °/\
(/o)
2.0
3.1
3.9
*
4.9
3.9
3.8
3.2
* Single measurement
Table 2.  Results from a Single-Laboratory Accuracy
            Evaluation of a Fluoride ISE
Fluoride
Concentration
(mg/L)
0.0250
0.0500
0.125
0.250
0.500
2.50
5.00
25.0
50.0
250.
500.
Fluoride
Detected
(mg/L)
0.066
0.085
0.164
0.31
0.51
2.43
5.0
25.2
53.
260.
530.
Fluoride
Recovery
(%)
264
170
131
125
103
97
101
101
105
105
107
Rel. Std.
Deviation
IV \
(/o)
19
11
10
7
2
2
2
2
3
3
3
                      9214 - 7
Revision 0
January 1995

-------
Table 3.  Mean Spike Recoveries for 12 Groundwater Samples and
                     6  EP  Leachate  Samples

     Spike         Groundwater             EP Leachate
  Fluoride   Fluoride   Rel.  Std.   Fluoride   Rel.  Std.
   Added     Recovered  Deviation   Recovered  Deviation
   (mg/L)    (percent)  (percent)   (percent)  (percent)
     0.5         92        5.3         100       20
     1.5         92        3.3          92        9.7
     3.5         93        3.0          91        5.5
     8.5         96        2.6          91        4.1
                           9214 - 8                       Revision  0
                                                          January  1995

-------
                  METHOD 9214

   POTENTIOMETRIC DETERMINATION OF  FLUORIDE
IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE
Start
                        J)
^
r
7.1.1 - 7.1.2
Calibrate Fluoride
ISE.
•^
r
7.2 Allow standards
to equilibrate to
room temperature.
>
r
7.3 Rinse
eletrodes.
>
r
7.4 Measure
concentration using
electrode meter
and calculate
concentration.
•^
i
7.5 Drain
reference electrode
and clean.
>
r
                   9214 - 9
                                 Revision 0
                                 January 1995

-------
                                  METHOD 9215

                    POTENTIOMETRIC DETERMINATION  OF  SULFIDE
        IN AQUEOUS SAMPLES AND DISTILLATES  WITH  ION-SELECTIVE  ELECTRODE
1.0   SCOPE AND APPLICATION

      1.1   This method can be used for measuring total sulfide in a distilled
sample.   The method is  meant to be  used  as an alternate  determinative step
following the distillation in either SW-846 Methods 9030 or 9031.

      1.2   This method  must  not be  used  for undistilled  samples  because  of
possible mercury and silver ion interferences.  Also, the ISE only responds to
free sulfide dianion and will  not  detect  sulfide  in complexes.  Therefore, this
method would provide low recoveries for undistilled samples.

      1.3   The method  detection limit is 1.0 mg/L.  Sulfide concentrations from
0.1 to 12,000 mg/L may  be measured.  However, when a linear calibration is used,
results less than 1 mg/L may be biased up to approximately 90 percent low.

2.0   SUMMARY OF METHOD

      2.1   The distillations  in  Methods  9030 and/or 9031 are performed, except
that the scrubber solution is  sulfide  anti-oxidant buffer (SAOB), with ascorbic
and  salicylic  acids added  as oxygen  scavengers.    The distillates  are then
analyzed potentiometrically using a sulfide  ion-selective  electrode  (ISE)  in
conjunction with a double-junction  reference  electrode  and  a pH meter with  an
expanded millivolt scale or an ISE meter capable of being calibrated directly in
terms of sulfide concentration.

      2.2   This method is an alternative to the iodometric titration procedure,
where a solution of thiosulfate is  standardized against the thiosulfate solution,
and the sulfide standard is standardized against the iodine solution/thiosulfate
solution  (which  requires  daily calibration  itself).   This  method  allows for
standardizing the sulfide calibration  standards  by  a potentiometric titration
with standardized silver nitrate using the sulfide ISE as the working electrode.
Silver  nitrate   solutions  are  stable when  stored  properly  and  are  easily
standardized against sodium chloride.

      2.3   The key to  acceptable  recoveries is the use  of the proper apparatus
and careful assembly of the distillation apparatus prior  to distillation.  If
these steps are not taken, low recoveries will result.

3.0   INTERFERENCES

      3.1   Since this  method  may  only be used for distillates, which will have
a pH greater than 12, there are no expected  interferences.  As shown in Table 1,
the data indicate that neither silver,  mercury, nor humic acid interfere with the
electrode.  (An interference would be  indicated by  a positive response.)

      3.2   The  effect of  sulfur(+4)  compounds  (which  interfere  with  the
titrimetric method)  in the  absorbing  solution  on  the sulfide  ISE  was  tested


                                   9215 - 1                       Revision 0
                                                                  January 1995

-------
(e.g.,  sulfite  or  thiosulfate).    The electrode  did not  respond to  either
compound.  A 110 mg/L sulfide solution  was  spiked with 100 mg/L sulfide and 100
mg/L  thiosulfate.   The  measured  concentrations for  the unspiked  and  spiked
samples were identical (114 mg/L).

      3.3   Temperature  changes  affect electrode  potentials.    Standards  and
samples must be equilibrated at the same temperature (+ 1°C).

      3.4   The user  should  be aware of the potential of  intereferences from
colloidal substances and that, if necessary, the samples  may be filtered.

4.0   APPARATUS AND MATERIALS

      4.1   pH/mV meter capable of reading to 0.1 mV or an  ISE meter.

      4.2   Sulfide ISE (Orion  9416 or equivalent) and double-junction reference
electrode (Orion 902000 or equivalent).

      4.3   Thermally isolated magnetic stirrer, Teflon®-coated  stir  bar,  and
stopwatch.

      4.4   Volumetric flask,  125 mL.

5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all reagents  shall conform to the specifications
of the Committee on Analytical  Reagents of  the American Chemical Society, where
such  specifications  are  available.   Other grades may be used, provided it is
first ascertained that the reagent is of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent water.   All  references to  water  in  this method  refer to
reagent water, as defined in Chapter One.

      5.3   Sulfide anti-oxidant buffer (SAOB).  Dissolve 80 g NaOH,  320 g sodium
salicylate and 72 g ascorbic acid in 1 L reagent water.   Prepare  fresh weekly.

      5.4   Standard sodium  chloride solution  (0.100N).  Dissolve  5.84  g NaCl
(dried for 2 hours at 140"C) in water and dilute to 1.00  L  with reagent water.

      5.5   Sodium  hydroxide (6N),  NaOH.    Dissolve  240  g  of  NaOH in  1  L of
reagent water.  Keep tightly closed.

      5.6   Potassium chromate indicator  solution.   Dissolve 50  g K2Cr04 in a
little reagent water.  Add  AgN03  solution  until  a  definite  red precipitate is
formed.  Let stand 12 hrs, filter, and dilute to 1  L.

      5.7   Standard silver nitrate  solution (0.10N).  Dissolve 16.99 g of AgN03
(dried for 2 hours at 150"C) in reagent water  and dilute to  1.00  L.  Store in a
brown bottle.  Standardize weekly against standard sodium chloride solution.

      5.8   Sodium  sulfide  nonahydrate,  Na2S  •  9H20.   For  the preparation of


                                   9215 - 2                       Revision 0
                                                                  January 1995

-------
sulfide standard solutions to be used for calibration curves.  Standards .must be
prepared between pH 9 and pH 11.  Protect standards from exposure to oxygen by
preparing them without headspace.  These  standards  are  unstable and should be
prepared daily.

      5.9   Sodium sulfide (2% and 10%), Na2S.

      5.10  Ammonia, NH3:   concentrated.

6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   All  samples must  have been  collected  using a sampling  plan that
addresses the considerations discussed in Chapter Nine of this manual.

      6.2   All original,  undistilled samples should be  handled as described in
Section 6 of Methods 9030  and/or 9031.  Samples should be stored at  4°C prior to
analysis.

      6.3   The distilled  samples should  be analyzed with  the  ISE  immediately
following distillation, distillates that are not analyzed immediately should be
stored in a sealed flask at 4°C.

7.0   PROCEDURE

      7.1   Replace the scrubber solution in Method 9030 or  9031 with 10.0 mL
SAOB solution and 40.0 mL reagent water.  Proceed to distill  samples as described
in those methods.

      7.2   Standardization of silver nitrate

            7.2.1 Add 10.00 mL of 0.100N NaCl  and 40 mL reagent water to a 125
      mL flask.   Adjust  pH  to 7-10  with  dilute NaOH  solution.   Add  1.0 mL
      potassium chromate  indicator.   Titrate with silver nitrate solution to a
      pinkish-yellow endpoint.  Be consistent with endpoint  recognition.  Repeat
      with a reagent blank  (water and indicator).   Calculate the normality of
      the silver nitrate  as follows:

            N AgN03 = (A  - B)   x  N(NaC1)
                            10.00 mL

            A = mL titration for NaCl
            B = mL titration for reagent blank

      7.3   Standardization of sulfide standards

            7.3.1 From the sodium sulfide salt,  prepare standards with nominal
      concentrations of 10, 100,  and  1,000 ppm sulfide in a matrix of 20% SAOB.
      Standardize each solution immediately prior to calibrating the ISE.  The
      standards may be calibrated by  iodometric  titration or by potentiometric
      titration as described below.

            7.3.2 The titration is monitored with the combination silver/sulfide
      electrode  (silver-coated  platinum   ring  sensing  electrode   with  a


                                   9215 - 3                        Revision 0
                                                                  January 1995

-------
      syilver/silver chloride reference electrode).   Prior to use the electrode
      is conditioned by soaking in 2% sodium sulfide for 5 minutes,  soaking in
      10% sodium sulfide until the  brownish  layer  becomes  black,  rinsing with
      reagent water, and cleaning with a  soft cloth.   After conditioning,  the
      electrode is connected to the pH/mV meter.   20  ml  of a sulfide standard
      (or suitable  quantity  to  get  accurate titration)  and  1  ml  concentrated
      ammonia are pipetted into a titration  vessel.  The electrode is inserted
      and the potential  recorded.  The sample is titrated with the standardized
      silver nitrate until  a potential of  100 mV is attained.  The potential is
      recorded after each titrant addition.  The equivalence point is determined
      from  the  first  derivative  of  the   titration  curve.     The  sulfide
      concentration is then calculated as  follows:

            sulfide (ppm)  = A  x  B  x  16,000 mg/eq
                                       C

            A = ml silver nitrate
            B = normality of silver nitrate
            C = ml of sulfide standard

      7.4   Calibration of Sulfide ISE and Meter

            7.4.1 Calibrate the sulfide ISE using  the  10,  100,  and  1,000 mg/L
      sulfide standards.  The standards must be freshly standardized.  Add 25.0
      ml of standard into a  50 ml beaker.   Add  a Teflon®-coated magnetic stir
      bar, place the beaker  on a magnetic stir plate,  and  stir at slow speed (no
      visible vortex).  Immerse the electrode tips  to  just above  the rotating
      stir bar.  If using  an ISE meter, calibrate the meter in terms of sulfide
      concentration following the manufacturer's instructions.  If using a pH/mV
      meter, record the  meter reading (mV) as soon as the  reading is stable, but
      in  no case  should  the  time  exceed  five  minutes  after immersing  the
      electrode  tips.    Prepare  a  calibration  curve  by  plotting  measured
      potential (mV) as a function  of the logarithm of sulfide concentration.
      The slope must be 54-60 mV  per decade  of  sulfide concentration.   If the
      slope  is not  acceptable,  the  ISE  may not  be  working  properly.    For
      corrective action, consult the ISE operating  manual.

      7.5   Allow samples  and standards to equilibrate to  room temperature prior
to analysis by ISE.

      7.6   Prior to and between analyses, rinse the electrodes thoroughly with
reagent water  and  gently  shake off excess water.   Low-level measurements are
faster  if  the  electrode tips are first  immersed  for  five minutes  in  reagent
water.

      7.7   Measurement of Sulfide in Distilled Samples

            7.7.1 Pour 25.0 ml of sample  into a 50  ml  beaker.   Add  a Teflon®-
      coated magnetic stir bar.  Place the beaker on a magnetic stir plate and
      stir at a slow speed (no visible vortex).   Immerse the electrode tips to
      just  above  the  rotating  stir  bar.    Record  the  meter  reading  (mV or
      concentration) as soon as the reading is stable, but in  no case should the
      time exceed five minutes after immersing the  electrode tips.  If reading
      mV, determine sulfide concentration from the  calibration curve.

                                   9215 -  4                       Revision 0
                                                                  January 1995

-------
      7.8   When analyses have been completed, rinse the electrodes thoroughly
and store them in a 1.0 mg/L sulfide standard  solution.   If the electrodes will
not be used more than  one day,  drain the reference electrode internal filling
solutions, rinse with reagent water, and store dry.

8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter One  and the  distillation methods  for specific
quality control procedures.

      8.2   Initial Calibration Verification standard (ICV):  After performing
the calibration step,  verify calibration  by  analyzing  an ICV.  The ICV contains
a known sulfide concentration at the mid-range  of the calibration standards and
is from an independent source.  ICV recovery  must be 90-110 percent.   If not, the
error source must  be found  and  corrected.   An acceptable ICV must be analyzed
prior to sample analysis.  The ICV also serves as a laboratory control sample.

      8.3   Continuing Calibration Verification standard (CCV):  After every 10
samples, and after the final sample, a CCV must be analyzed.  The CCV contains
a known sulfide concentration at mid-calibration range.  CCV recovery must be 90-
110 percent.  If not,  the error  source  must  be  found and corrected.  If the ISE
calibration has changed,  all  samples analyzed since the last  acceptable CCV must
be re-analyzed.

      8.4   Reagent blank:  After the ICV and after every CCV, a reagent blank
must be analyzed.  A reagent blank is reagent water mixed 100:1 with SAOB.  The
indicated reagent blank concentration must be less than 0.05 mg/L sulfide.  If
not, the contamination  source must be found and corrected. All samples analyzed
since the last acceptable reagent blank must be re-analyzed.

      8.5   Matrix spike:  Follow the matrix  spike protocols  presented in Chapter
One.  The  spike concentration must be 10 times the detection  limit and the volume
added must be negligible (less than or equal to one-thousandth the sample aliquot
volume).   Spike recovery  must  be  75-125  percent.    If not, samples  must  be
analyzed by the method of standard additions.

      8.6   The sulfide calibration standards  may degrade by  more than 10% from
day-to-day.  The standards must  be  standardized daily before  use (by titration)
and checked throughout the day if used as QC samples.

9.0   METHOD PERFORMANCE

      9.1   The sulfide ISE was calibrated with  100 and 1,000 mg/L standards, and
a series of sulfide standards was  analyzed as unknowns.   The  results are listed
in Table 2.  As shown, recoveries ranged from 76-124% over the range 0.25-12,000
mg/L sulfide.   This indicates that  there  is  no  practical  difference between the
true and observed values for sulfide over this concentration range.

      9.2   Three acid-soluble sulfide samples were prepared (low,  medium, and
high; 25,  100,  and  1,000 mg/L).  Triplicate  measurements  of each were performed
in the following order:   medium,  low,  high, high, low,  medium.  The data are
provided in Table 3.

      9.3   Precision estimates were  calculated  from the average  percent RSD

                                   9215 - 5                       Revision 0
                                                                  January 1995

-------
taken within  each  block of triplicate measurements  in  Table 3.   The  CV% was
calculated as 2.0 to 10.

      9.4   Accuracy estimates from average percent recovery reported in Table
3 were calculated to be 75-105%.

      9.5   To test the  efficiency of the SAOB as a  scrubber  solution, three
standards  (1,  10,  and  40  mg/L)  were distilled  into  SAOB  scrubber  and the
resulting sulfide concentration measured.  The  results  are  listed  in Table 4.
As seen in the table,  excellent recoveries  are obtained  using the SAOB scrubber
solution.  The one low recovery for the 40 mg/L standard is most likely due to
incomplete sparging of oxygen from the system prior to distillation.

10.0  REFERENCES

      1.    Franson, Mary Ann H.,  Ed.  Standard Methods for the Examination of
Water  and Wastewater,  18th  Edition.    American  Public  Health  Association,
Washington, DC, 1992.

      2.    Model  94-16  Silver/Sulfide Electrode  Instruction  Manual.   Orion
Research, Inc., Boston, MA, 1986.

      3.    Hillerman, D.C., Nowinski, P.   "Modification  of Methods 9030  and 9031
for  the  Analysis  of Sulfide  by Specific  Ion Electrode".   U.S.  Environmental
Protection Agency, EMSL-LV.  EPA/600/4-90/024.   September 1990.
                                   9215 - 6                       Revision 0
                                                                  January 1995

-------
        Table 1.  Interference Study Results
Interferant
Ag+ (20 mg/L)
Ag+ + NH4+ (20 mg/L
each)
Hg+ (20 mg/L)
Humic acid (100 mg/L)
Humic acid + sulfide
(100 mg/L each)
Sulfide Concentration (mg/L)
DI matrix
<1
<1
<1
2.1
114
Tap water matrix
<1
<1
<1
1.6
87,8
* The humic acid contains 2-3 mg/L sulfide as determined by Method 9030.
Table 2.  Results From a Single-Laboratory Accuracy
             Evaluation  of  a  Sulfide  ISE
Sulfide Sulfide
Concentration Detected
(mg/L) (mg/L)
0.10
0.25
0.50
1.00
1.09
5.00
27.4
55.0
110
591
1,183
6,000
12,000
0.01
0.19
0.47
0.99
1.10
4.94
26.8
52.3
109
607
1,157
6,028
14,850
Sulfide
Recovery
(percent)
10
76
94
99
101
99
98
95
99
103
98
100
124
Rel. Std.
Deviation
(percent)

-
12.7
9.3
.
3.4
-
-
-
-
0.4
2.0
2.2
                      9215 - 7
Revision 0
January 1995

-------
                                                                         i
Table 3.  Precision and Accuracy Data for the Sulfide ISE
True
(mg/L)
23.6
118.1
1,503
Measured
(mg/L)
19.2
114.8
1,455
Std. Dev.
(mg/L)
0.81
2.2
26
RSD (%)
4.2
1.9
1.8
Recovery
(%)
82.5
97.2
96.8
n = 6 for all samples
 Table 4.  Recovery of Hydrogen Sulfide in SAOB Solution
Sulfide (mg/L)
1
10
40
% Recovery
91.0
89.8
86.7
96.6
100
96.0
69.3
98.9
89.2
                        9215  - 8
Revision 0
January 1995

-------
                   METHOD 9215
     POTENTIOMETRIC DETERMINATION OF  SULFIDE
IN DISTILLED SAMPLES WITH ION-SELECTIVE ELECTRODE

r
7.1 Add 10 mL of SAOB
solution and 40 mL of
reagent water; proceed
to distill sample.
>
r
7.2 Standardize
silver nitrate.
>
r
7.3 Standardize sulfide
standards.
1
r
7.4 Calibrate sulfide ISE.
>
f
7.5 Allow samples &
standards to equilibrate
to room temperature.
>
r
7.6 Rinse electrodes.
>
r
7.7 Take measurement &
determine concentration.
>
r
7.8 Drain reference
electrode and rinse.
^
r
                     9215 - 9
Revision 0
January 1995

-------
                            CHAPTER  SIX

                            PROPERTIES
The following methods are found in Chapter Six:
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
1030:
1120:
1312:
1320:
1330A:
9041A:
9045C:
9050A:
9080:
      Method 9081:
      Method 9090A:
      Method 9095A:
      Method 9096:
            Appendix A:
      Method 9100:

      Method 9310:
      Method 9315:
Ignitability of Solids
Dermal Corrosion
Synthetic Precipitation Leaching Procedure
Multiple Extraction Procedure
Extraction Procedure for Oily Wastes
pH Paper Method
Soil and Waste pH
Specific Conductance
Cation-Exchange  Capacity   of   Soils  (Ammonium
Acetate)
Cation-Exchange Capacity of Soils (Sodium Acetate)
Compatibility Test for Wastes and Membrane Liners
Paint Filter Liquids Test
Liquid Release Test (LRT) Procedure
      Liquid Release Test Pre-Test
Saturated  .Hydraulic    Conductivity,   Saturated
Leachate Conductivity, and Intrinsic Permeability
Gross Alpha and Gross Beta
Alpha-Emitting Radium Isotopes
                              SIX  -  1
                                               Revision 2
                                               January 1995

-------
                                  METHOD 1030

                            IGNITABILITY OF SOLIDS
1.0  SCOPE AND APPLICATION
      1.1  This method is suitable for the determination of the ignitability of
solids and is appropriate for pastes, granular materials,  solids that  can be cut
into strips,  and  powdery substances.  This method may  be  used  to meet certain
regulatory applications but is  not required for determining if a waste passes or
fails the  characteristic of ignitability per  the  RCRA definition.   If  it is
impractical  to  perform the test  because  of  the physical form of the sample,
generator knowledge should be used to determine the  ignitability hazard posed by
the material.

2.0  SUMMARY OF METHOD

      2.1  In a preliminary test, the test material is formed into an unbroken
strip or powder train  250 mm in length. An ignition source is  applied  to one end
of the test material to determine whether combustion will  propagate along 200 mm
of the strip within a specified time period.   Materials that propagate burning
along a 200  mm  strip  within the  specified time period are then subjected to a
burning rate test.   Materials that do not  ignite or  propagate  combustion as
described above do not require further testing.  In the burning rate test, the
burning time  is measured  over  a  distance  of  100 mm and the  rate  of burning is
determined. The test method described here  is based on the  test procedure adopted
by the Department of Transportation  from the United Nations regulations for the
international transportation of dangerous goods and is contained in Appendix E
to Part 173  of 49 CFR.

3.0  INTERFERENCES

      3.1   In  laboratory tests the burning rate of duplicate  runs is usually
repeatable to within 10%.  However, large differences in burning rates may occur
if experimental conditions are not held constant.  Variation in airflow rates,
particle  size,  and moisture  content   of  the  test  material  will  affect  test
results.   Therefore,  at  least triplicate  determinations of the  burning  rate
should be conducted.

      3.2  Particle size of test  material can affect not only the burning rate,
but also the ignition  of the material.   Therefore, the  particle  size of the test
material should be the same for  each  test run.   The  particle size of the test
material should be reported in a simple descriptive format (e.g., fine powder,
sand, coarse granular).

      3.3  Temperature of some test material  such as sulfur powder affects the
burning rate.   For reproducible results, all  tests  should be  performed  at
approximately  the  same  initial  temperature  (ambient  room  or  laboratory
temperature).


                                  1030 -  1                       Revision 0
                                                                  January 1995

-------
      3.4   All  tests must  be carried  out  inside a  fume  hood with  the  test
apparatus situated  perpendicular  (90°)  to  the direction of  airflow.   Airflow
parallel (0°)  to the test apparatus results in non-reproducible burning rates.

      3.5  The rate of airflow through the fume hood affects the burning rate.
Too high an airflow distorts the flame and retards its horizontal  propagation.
The optimum airflow appears to be in the range of 0.7-1 meter per second.

      3.6  Materials that are moisture sensitive (i.e.,  readily  absorb  moisture
from air) should  be tested  as  quickly  as possible after removal from the sample
container.  All  materials should be tested as received by the laboratory.

4.0  APPARATUS AND MATERIALS

      4.1   Low-heat conducting,  non-combustible,  impervious  ceramic  tile  or
equivalent material, of approximate dimension of 25 cm x 25 cm x 2.5 cm (the tile
must be at least 25 cm in length to support a 250 mm test sample).

      4.2   High  temperature  marker  or equivalent  making  device for  marking
ceramic plates.

      4.3  Powder Train Mold  (see  Figure  1)  for molding powdered  and granular
materials for the  burn rate test.  The material  of construction  can be aluminum,
brass, or stainless  steel.  The mold  is 250  mm in  length and has  a  triangular
cross-section, with a width of 20  mm,  and  a depth of 10 mm as measured from the
bottom of the triangular  opening to where  the sides meet. On both sides of the
mold, in  the  longitudinal  direction,  two  metal sheets  are  mounted  as lateral
limitations which  extend 2  mm beyond the upper edge  of the triangular cross-
section.  This device can be fabricated by most machine shops. The complete burn
rate apparatus is  available from:  Associated Design and Manufacturing Co.; 814
N. Henry Street;  Alexandria, Virginia 22314.

      4.4  A Bunsen  (propane gas and air)  burner  with a minimum diameter of 5
mm capable of attaining a temperature of at least 1,000°C.

      4.5  Stop watch

      4.6  Thermocouple to measure the temperature of the gas flame.

      4.7 Thermometer to  measure  initial  temperature  of material  (i.e.,  room
temperature).

       4.8 Anemometer to measure airflow in the fume hood.

5.0  REAGENTS

      5.1  No special reagents are required to conduct this test.
                                   1030 -   2                      Revision 0
                                                                  January 1995

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All  samples  are  tested on  as-received  basis  unless  requested
otherwise.  No sample preservation  is required, but sample containers should be
completely filled and tightly sealed to preserve sample integrity.

      6.2   Samples should be tested as  soon as possible after removal from the
sample container (i.e., samples should  not  be allowed to dry or absorb moisture
for excessive  periods  or to loose volatiles).  Samples that  are chilled or cooled
upon receipt to the laboratory should be allowed to equilibrate to the ambient
laboratory temperature in the sample container.

7.0  PROCEDURE

            SAFETY; Prior to  starting the preliminary test, all sample materials
            must  be   tested  to  determine  if that  material  is  explosive  or
            extremely flammable.  Use a very small portion of material  (1 gram
            or  less).     If  the   sample   displays   explosivity  or  extreme
            flammability, do not conduct this test.

      7.1  Preliminary Screening Test

            7.1.1  The preliminary ignitability  test is conducted on all  waste
      materials.   On  a clean,  impervious ceramic tile  (Section  4.1),  clearly
      mark a 250 mm long  test  path.  Make  another mark at exactly  200  mm from
      the start of the sample path.

            7.1.2   Prepare  the  test material  in its  "as  received"  form  by
      forming  an unbroken strip or powder train  of sample  250 mm long by 20 mm
      wide by  10 mm high on the ceramic  tile. Use  the mold to form the material
      as in 7.2.3 if appropriate.

            7.1.3 Place the ceramic tile with the loaded sample in a fume hood
      about 20  cm  (~8 inches)  from  the front of the  hood  and in  an  area  of
      laminar  airflow. Position  the  sample  perpendicular  to  the  airflow. (See
      Figure 2)  The  airflow across the  perpendicular axis of the sample should
      sufficient to prevent fumes from escaping  into the laboratory and should
      not be varied during the test.  The air velocity should be approximately
      0.7 meters/second.  Measure the air velocity by an anemometer.

            7.1.4  Light  the Bunsen  burner  and  adjust  the height of the  flame
      (6.5 to  7.5 cm) by adjusting the propane gas and air flows.  Measure the
      temperature  of  the  flame (tip  of the flame)   by  a thermocouple.  The
      temperature of the flame must be at least  1000°C.

            7.1.5  Apply the tip of the flame to one end of the sample  strip .
      The test period will depend on the sample  matrix as  follows:

            7.1.6   If the waste  is  non-metallic, hold the  flame tip on  the
      sample strip until  the sample  ignites or for a maximum of  2  minutes.  If


                                   1030  -  3                       Revision 0
                                                                  January 1995

-------
      combustion occurs, begin  timing  with  a stop watch and  note  whether the
      combustion propagates  up  to the  200  mm mark  within  the 2  minute  test
      period.

            7.1.7  If the waste  is a metal  or metal-alloy powder, hold the flame
      tip on the sample strip  until  the sample ignites or for  a maximum  of 5
      minutes.   If combustion occurs,  begin  timing with a stop  watch  and  note
      whether the combustion propagates up to the  200 mm mark within  the 20
      minute test period.

            7.1.8  If the waste  does not ignite and propagate combustion either
      by burning with open flame or by smoldering along 200  mm of sample strip
      within the  2  minute  test period (or  20 minute  test  period for metal
      powders),  the  waste is not  considered flammable and no further testing is
      required.   If  the waste  propagates  burning of 200 mm of  the test strip
      within the 2 minute test  period  (20 minute  test  period  for metals), the
      material  must be evaluated by the burning rate test  (Section  7.2).

7.2  Burning Rate Test

            7.2.1  The preparation of  the test  sample for the burning rate test
      will depend on  the physical  characteristics of the waste.   Wastes  that
      exist in  a powdered or granular  state  are  molded  in a  powder train  mold
      shown in  Figure 1. Pasty materials are formed into a rope 250 mm in length
      with a cross-section  of   1  cm2.   All  tests for  the  burn rate  test are
      performed on clean,   ambient temperature, ceramic plates.

            7.2.2  On a clean,  impervious ceramic tile (Section 4.1),  clearly
      mark a 250 mm long test path.  Make  two additional  timing marks  at 80 mm
      and 180 mm from the start of the  sample  path.   The  distance  between the
      two marks (100 mm) will  be used to calculate the rate  of burn in Section
      7.2.9.

            7.2.3 Tighten the side plates on the  mold.  For powdered or granular
      materials:  Place the mold  on the base  plate.  Pour  the  material  to  fill
      the triangular cross section of the  mold loosely.

            7.2.4   Drop the unit  from a height  of  2 cm onto a solid surface
      three times to  settle  the powder.   Remove  the side supports.   Lift the
      mold  off  the  base plate.   Place a  clean  ceramic  test  plate  with the
      appropriate timing marks  (Section 7.2.2) face down on  top of the mold.
      Invert the setup and remove  the mold.

            7.2.5  Pasty wastes are prepared by spreading the waste on a marked
      ceramic tile (Section 7.2.2) in the  form of a rope 250 mm in  length with
      a cross-section of 1  cm2.

            7.2.6   Place the ceramic  tile with the  loaded  sample  prepared in
      Sections  7.2.3 or 7.2.5  in  a fume hood  about  20  cm  (~8 inches)  from the
      front of the hood and  in  an  area  of laminar  airflow.  Position the sample


                                   1030 -  4                      Revision 0
                                                                  January 1995

-------
      perpendicular  to  the airflow.  (See  Figure 2)   The airflow  across the
      perpendicular axis of the sample should sufficient to prevent fumes from
      escaping into  the  laboratory and should not be  varied  during the test.
      The air velocity should be approximately 0.7 meters/second.  Measure the
      air velocity with an anemometer.

            7.2.7  Light the  Bunsen  burner  and  adjust  the height of the flame
      (6.5 to 7.5 cm) by adjusting the propane gas and air flows.  Measure the
      temperature  of the  flame (tip  of the  flame)   by  a thermocouple.  The
      temperature of the flame must be at least 1000°C.

            7.2.8  Apply the tip of the flame  to  one end of the sample strip to
      ignite the test strip as described in Section 7.1.6 and 7.1.7.

            7.2.9  When the test strip or powder train has burned up to the 80
      mm time marker,  begin  timing the rate of  combustion with  a stop watch.
      Stop the  timer when the  burned strip  reaches  the  180 mm time marker.
      Record the amount of time  (in  seconds)  required  to  burn the 100 mm test
      strip.   Calculate the rate of burning by dividing the length of the burn
      test strip (100 mm) by the total time (seconds).   Results of the burn rate
      test should be reported  in mm/sec.  Wastes  that have a rate  of burning of
      more than 2.2 mm/sec  (or burn time of less than  45 seconds for  100 mm) are
      considered to  have  a positive result for  ignitability  according  to DOT
      regulations.   For metals,  this time is 10 minutes or less for 100 mm (or
      a burn rate of more than 0.17 mm/sec).

            7.2.10  Report and Calculation Section

Test Material Information

Source of Material: e.g., Company,  operation or process
Description of material:  e.g.,  powder or paste, metallic or non-metallic
Particle size:  e.g., fine powder,  granular, sand, etc.
Preliminary Burning Time: 	 seconds.

Test Conditions

Date of Test:
Temperature of test material   (°C):
Air velocity through fume hood (m/s):
                                   1030  -  5                       Revision 0
                                                                  January 1995

-------
Ignitability Test Data
Test
Number
1
2
3
Time (sec)
elapsed between
application of
flame and start
of ignition



Burning
time over
100 mm
(sec)



Burning
Rate
(mm/sec)



Comments



8.0  QUALITY CONTROL

      8.1   All  tests  must  be performed  on  a  clean ceramic  plate  at  room
temperature.  All samples must have  been  collected  using  a  sampling plan that
addresses the considerations discussed in Chapter Nine of this manual.

      8.2  All replicate runs must be at the same initial  temperature (ambient
laboratory temperature).

      8.3  All  replicate  tests must be run at  approximately  the  same airflow
through the fume hood.

      8.4  Only materials  of the  same particle size distribution should be used
for all replicate tests.

      8.5  The burn  rate test must be conducted  in  triplicate if the preliminary
screening test is positive.  Any burn  rate  for non-metallic samples that exceeds
2.2 mm/sec (or a burn time of less  than 45 seconds for 100 mm) is considered to
have a positive  result.   For  metals,  a burn rate  of more  than 0.17 mm/sec (or
burn time of less than 10 minutes for 100 mm)  is considered to have a positive
result.

9.0  METHOD PERFORMANCE

      9.1   An independent laboratory validation was conducted  on the robustness
of the burn  rate test  procedure.  The materials  selected  for this evaluation
included:

1.  A 50/50 mixture of metallic silicon and lead dioxide (PB02)
2.  Excelsior
3.  Dextrin (yellow powder)
4.  Sulfur (fine yellow powder)
5.  Aluminum metal  (coarse)
6.  Magnesium metal (coarse)
                                   1030 -   6
Revision 0
January 1995

-------
7.  Polyethylene high density (granular)
8.  Polyethylene low density (fluffy white powder)
9.  Scott fertilizer (32-3-10:N-P-K)
10. JP-4 contaminated soil (approximately 5000 ppm)

      Of these materials,  the 50/50 mixture of metallic silicon and lead dioxide
(PB02),  elemental sulfur,  and   excelsior were  considered to give  a positive
ignitability result under the conditions of the test.  The remaining materials
gave negative (nonflammable) results under the conditions of the test. Several
test variables  including  ignition source, ambient  temperature,  and apparatus
orientation, were studied using these materials.  Partial  results of this study
are summarized in Table 1.

                   Table 1  Test Variables for Ignitability
Material
Tested
50% Metallic
Silicon and 50%
Lead IV Oxide
Excelsior
(wood shavings)
Test
Number
1
2
3
4
1
2
3
4
Variable
combination1
ABC
Abe
aBc
abC
ABC
Abe
aBc
abC
Burn Time
over 100 mm
(sec)
0.84
0.50
0.69
0.65
13.45
9.14
13.37
13.59
Burn Rate
(mm/sec)
119
200
145
154
7.43
10.9
7.47
7.36
1where:
A-flame ignition
a-hot wire ignition source
B-ambient temperature of 20°C
b-ambient temperature of 100'C
C-orientation of test apparatus of 90* to air flow
c-orientation of test apparatus of 0° to air flow
      9.2  In another evaluation of the DOT burn rate test, potentially ignitable
finishing wastes  from the  furniture  industry were  collected and  tested  for
burning rates.    Each waste was tested in triplicate  to establish a mean value
for the burning rate.  The  results  for  the  flammable wastes  are  summarized in
Table 2.
                                   1030  -  7
Revision 0
January 1995

-------
                  Table 2-Burning Rates for Ignitable Wastes
Sample No.
A2
J2
U
K
H
F
P
0
Q
Description of Waste
Segregated Lacquer Dust
Segregated Lacquer Dust
Segregated Lacquer Dust
Consolidated Lacquer Dust
Catalyzed Lacquer Dust
Water Based Lacquer Dust
Booth Coat-Stain Overspray
Pallet Covered Cardboard
Pallet Covered Cardboard
Burn Time over 100
mm (sec)
4.7
4.6
8.6
6.0
6.7
19.4
12.5
11.1
12.3
Burn
Rate
(mm/sec)
21.3
21.7
11.6
16.7
14.9
5.15
8.0
9.0
8.13
      9.3  In  order to evaluate the ruggedness of the DOT burn rate test, select
ignitable finishing wastes were split and tested by a  state  laboratory and an
independent contract  laboratory.   The results of this comparison are summarized
in Table 3.

                       Table 3-Comparison of Burn Rates
Sample No.
Al
Jl
12
Description of
Waste
Segregated Lacquer
Dust
Segregated Lacquer
Dust
Booth Coat-Glaze
Overspray
Mean Burn Time Over 100 mm in
Seconds
State
Laboratory
4.7
4.6
O1
Contract
Laboratory
5
4.3
O1
1waste was found to be nonflammable under conditions of the test.
                                   1030  -  8
Revision 0
January 1995

-------
10.0  REFERENCES
      1.           "Test Methods for Readily  Combustible  Solids.   Burning Rate
                  Test."   (14.2.2.5).   Recommendations  on  the  Transport  of
                  Dangerous Goods.   Fifth  Revised Edition.  United Nations, New
                  York.  1988.

      2.           DOT Regulation.   Appendix E to Part 173 of 49 CFR, Chapter 1
                  (12-31-91 Edition),   pp. 597-598.

      3.           Flammability (solids).   Method A.10.  Official Journal of the
                  European Communities.  9/19/84.   No.  L251/63.

      4.           "Validation of Ignitability Method For Solids" Foster Wheeler
                  Enviresponse, Inc.,   Edison NJ.,  Submitted to the Office  of
                  Solid Waste, US  EPA, February 1994.

      5.           Internal Report,   (AMFA  Report) North  Carolina  Department  of
                  Environmental Health and Natural  Resources. (Bill  Hamner)
                                   1030  -  9                       Revision 0
                                                                  January 1995

-------
                              Figure 1

                         Powder Train Mold
(A)    Cross-section of 250 mm long mould
                            1030 -   10
Revision  0
January  1995

-------
              Figure 2

Test Apparatus Position in Fume Hood
T
8"
1

i 	 M


t
	 a- 	 '
AIR FLOW
HOOD SASH °-7 m/s
  SAMPLE 90* TO AIR FLOW
               1030  -   11
Revision 0
January 1995

-------
                             METHOD  1030
                      IGNITABILITY OF  SOLIDS
                                      i
               Nonmetallio
7.1.6 Apply flame to
  test strip, hold
  for 2 minutes.
>
1
7.1 1 Perform preliminary
screening test.
^
r
71.1-7.12 Mark
ceramic plate,
prepare test strip
or powder train.
1
r
7.1 3 Position apparatus
in fume hood 90°
to air flow.
^
r
7 1 4 Light flame and
measure temperature.
                                              Metallic
7.1.7 Apply flame to
  test strip, hold
  for 5 minutes.
 Nonflammable Solid
                                                               No


                                                       Nonflammable Solid
                              1030 -   12
                Revision  0
                January  1995

-------
                           METHOD 1030
                   TGNITABILITV  OF  SOLIDS
7 2.3 - 7.2.4 Pour waste
   into the mold, drop
   three times,  invert
   on marked ceramic
        plate.
7 2.5 Spread pasty
 waste on marked
  ceramic plate,
 250 mm x  1 cm2.
                      7.2.6 Position test sample
                        in a fume hood 90 to
                              atr flow.
                         7.2 7 Light bunsen
                        burner, adj. flame and
                        measure  temperature.
                       7.2 8 Apply flame to end
                         of the strip to ignite.
                      7.2.9 Note time (seconds)
                        needed for the flame to
                       travel 100 mm distance.
                         Calculate burn rate.
                               1030  -   13
                           Revision  0
                           January  1995

-------
                                  METHOD 1120

                               DERMAL CORROSION
1.0   SCOPE AND APPLICATION

      1.1  The dermal corrosion  assay  system  is  an  in vitro test method which
determines the corrosive potential  of a substance toward human skin.  The method
is  simple,  rapid, accurate,  and may  be  applied  to both  solids,  liquids and
emulsions.  The  liquids may be  aqueous or non-aqueous.   Solids can be water-
soluble  or  non-soluble.   The  samples  may  be pure  chemicals,  dilutions,
formulations, or  waste.   No prior treatment of the  sample is required.   This
method may be used to meet certain regulatory applications but  is not required
for determining if a  waste passes or fails the characteristic of  corrosivit.y per
the RCRA definition.
2.0   SUMMARY OF METHOD

      2.1  The assay system is an in vitro  test method which is composed of two
components, a synthetic macromolecular biobarrier  and a Chemical Detection System
(CDS).   Test  samples  are  applied  on top  of the macromolecular biobarrier.
Corrosive  samples  are  able to  disrupt  the  macromolecular  structure  of the
biobarrier.   A color change  in the  CDS,  located beneath the  biobarrier,  is
detected visually and indicates that the  test  sample has altered the biobarrier
sufficiently to allow its passage through the  full thickness of the biobarrier.
The time it takes a sample to disrupt the biobarrier is inversely proportional
to the degree of corrosivity  of the sample  -  the longer it takes to observe a
color change, the less corrosive the substance is.  Noncorrosive samples do not
disrupt the  biobarrier,  or disrupt the  biobarrier after  a predetermined time
period (see Section 2.4).

           Corrosive  samples  may  be  placed  into  three different  classes  of
corrosivity, established by the time required for  the sample to break through the
biobarrier.  These classes are called Packing Groups by the U.S. Department of
Transportation (DOT).   Packing  Groups  are  assigned  according to the degree of
danger presented  by  the corrosive  material;  Packing  Group I  indicates great
danger; Packing Group II, medium danger;  Packing Group III, minor danger.  For
consistency,  these  same definitions  are used  for  this  test  method  and are
referred to as Group I,  Group II, and Group III.

      2.2  Prior to performing the assay,  samples  are pre-qualified to establish
their compatibility with  the  assay system.   The sample is placed  in  a small
amount of CDS fluid.   If any detectable change occurs in the CDS, the sample is
qualified and can be  analyzed  by the test.   If a  sample is non-qualified, it is
incompatible with the CDS and must be tested by another method.

      2.3  Test samples  are classified  into categories by the  screening test
which is  supplied with the  assay  kit.  The category that  a sample is assigned to
will  determine how the Groups  will  be  assigned.   Test samples are classified by
pH changes produced in two well-defined buffers  - one  designed  to buffer acids
                                   1120 - 1                       Revision 0
                                                                  January 1995

-------
and another  that buffers bases.   These buffers are  supplied as part  of the
screening test.  Four different categories are defined as follows:

           2.3.1  Category AT  substances produce a large change in pH when they
      are added to the acid  buffer.   This change in pH is indicated by a strong
      color change of the acid buffer solution.

           2.3.2  Category B,  substances produce a large change in pH when they
      are added to the base  buffer.   This change in pH is indicated by a strong
      color change of the base buffer solution.

           2.3.3  Category A2 substances produce little or  no pH changes when
      added to the acid buffer,  and  therefore,  little or no color change in the
      buffer solution is observed.

           2.3.4  Category B2 substances produce little or  no pH changes when
      added to the base buffer,  and  therefore,  little or no color change in the
      buffer solution is observed.

      2.4  Groups are assigned  in the  assay  system by  taking  into account the
category that  is  assigned to  a sample by the  screening test,  and  the time it
takes to detect  a color change  in  the  CDS  in  the  assay.  Category ^  and B,
samples are assigned to Group I if a color change  is  observed between zero and
three minutes, to Group  II if a color change is  observed after  three minutes and
up to one hour, and to Group III if a  color change is observed after one hour and
up to four  hours.   If  no color  change occurs  in  four hours,  the  chemical  is
classified as Noncorrosive.

           Category A2 and B2 samples are assigned to Group I  if a color change
is observed between zero and  three  minutes,  to Group II if a color  change is
observed after three minutes and up to 30 minutes,  and to Group III if a color
change is observed after 30 minutes  and  up to  45 minutes.   If no color change
occurs in 45 minutes,  the chemical is classified as  Noncorrosive.


3.0   INTERFERENCES

      3.1  The  test  is  not subject  to interference  from color,  turbidity,
colloidal matter or high salinity.

      3.2  The Pre-qualification Test,  the  Screening Test and  the Assay must be
performed at room temperature.  The samples must also be at room temperature (17-
25°C).


4.0  APPARATUS AND MATERIALS

     4.1   Corrositex Assay Kit (InVitro International,  16632 Millikan Avenue,
Irvine,  CA  92714).  The following  three items are supplied in the Corrositex
Assay Kit:

           4.1.1  Four racks holding seven vials with black caps.
                                   1120 - 2                       Revision 0
                                                                  January 1995

-------
           4.1.2  One tray of 24 membrane discs.

           4.1.3  Four data sheets (color charts).

     4.2   Combination hot plate/stir plate or equivalent - able to heat to 75°C.
Stirring speed should be adjustable.

     4.3   Digital thermometer - able to read to 75°C.

     4.4   Timers (6) - able to measure hours, minutes and seconds.

     4.5   Repeat pipettor -  this pipet is different than the  pipet  specified in
Section 4.12.  Delivers  200 /*L repeatedly, without refilling  between individual
deliveries.

     4.6   2.5 ml combitip for repeat pipettor.

     4.7   Lab Industries or equivalent sample pipettor - a positive displacement
pipettor useful when pipetting viscous samples.

     4.8   Pipet tips for Lab Industries, or equivalent, pipettor.

     4.9   Test tubes

     4.10  Balance - capable of weighing 100 mg accurately.

     4.11  Spatula - capable of transferring 0.1 - 0.5 g.

     4.12  Pipets -  microliter, with  disposable tips.  Should  be  able to measure
100  L accurately.

     4.13  Tweezers.

     4.14  Permanent marker pens.

     4.15  Plastic wrap.


5.0   REAGENTS

      5.1  All reagents listed below are provided  in  the Corrositex Assay Kit
except for the  positive  and  negative controls mentioned in  Section  5.7.   The
Corrositex Assay  Kit  is  available from  InVitro  International,  16632 Millikan
Avenue, Irvine, CA  92714.

      5.2  Chemical  Detection System (CDS).

      5.3  Screening test buffer solutions.

      5.4  Confirmation Test Solution.

      5.5  One gram of the biobarrier matrix and a microstir  bar.
                                   1120 - 3                       Revision 0
                                                                  January 1995

-------
      5.6  10 ml of biobarrier diluent.

      5.7  Positive and negative controls, if desired, for GLP purposes.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1  Appropriate  precautions  should be  taken  for  handling  potentially
corrosive substances such as wearing gloves and having proper eye protection.

      6.2  Samples should be analyzed as soon as possible after collection.


7.0   PROCEDURE

      7.1  Follow  the  established   laboratory procedures  for  working  with
hazardous test samples.  Wear lab coat, gloves and safety glasses when working
with any potentially corrosive material.

      7.2  Pre-Qualification Test

           7.2.1  Add 100 mg or 150 pi of sample to 1.0 mL of CDS in duplicate
      test tubes.

                  7.2.1.1  Sample qualifies if there is a color reaction within
           5 minutes: proceed with assay.

                  7.2.1.2  If  no  reaction  is  observed,  the  sample  is  non-
           qualified.  Seek other methods to determine corrosivity.

      7.3  Screening Test

           7.3.1  Liquid samples

                  7.3.1.1  Add 150  L of sample to Test Tubes 1 and 2. Cap the
           test  tubes  and shake  vigorously  for 10 seconds.   Read  the color
           change of the mixture within one minute.  If the sample is immiscible
           in the solution, wait  one  minute  and then  read  the color change  at
           the interface.

                  7.3.1.2  Assign the  category.    If  an  intense color change
           (similar  to  the Category  B,  color chart)   is  observed   in  Tube  1,
           assign  the  sample  to  Category  B,.    If  an  intense color change
           (similar  to  the Category  A,  color chart)   is  observed   in  Tube  2,
           assign the  sample  to  Category A,.   Proceed to the  next  step if  no
           intense color change is observed in Test Tubes 1 and 2.

                  7.3.1.3  Confirmation test. Add  two drops of the Confirmation
           Test Solution to Test Tube  1.  Cap the test  tube and shake vigorously
           for 5  seconds.   If the color of  the solution  changes to yellow  or
           gray  (similar  to  the  Category A2 color chart at the  bottom of the
           protocol  sheet)  the sample is  classified  as  Category A2.   If the
           color  of  the solution changes  to purple  or  blue  (similar  to the


                                   1120 - 4                       Revision 0
                                                                  January 1995

-------
     Category B2 color  chart at the  bottom  of the  protocol  sheet) the
     sample is classified as Category B2.

     7.3.2  Solid samples

            7.3.2.1  Add 100 mg of sample  to  Test Tubes  1 and 2.  Cap the
     test tubes and shake vigorously for one  minute.  Wait another minute
     and read the color  change of the mixture.  If the sample is insoluble
     in the  solution,  allow the  mixture  to  settle  and read  the  color
     change at the interface of the solution and the solid.

            7.3.2.2  Assign  the  category.    If an  intense  color change
     (similar to the Category  B.,  color  chart) is  observed in  Tube  1,
     assign  the  sample  to  Category  B,.    If an  intense  color change
     (similar to the Category  AT  color  chart) is  observed in  Tube  2,
     assign the  sample  to  Category Av   Proceed to  the next  step  if  no
     intense color change is observed in  Test Tubes 1 and 2.

            7.3.2.3  Confirmation test.  Add two drops of the Confirmation
     Test Solution  to Test Tube 1.  Cap the test tube  and shake vigorously
     for 5 seconds.   If the color of the solution changes  to  yellow  or
     gray  (similar to the  Category A2 color chart at the bottom of the
     protocol sheet) the sample is  classified as  Category A2.   If the
     color of  the  solution  changes  to purple or  blue  (similar  to the
     Category B2 color  chart at the  bottom  of the  protocol  sheet) the
     sample is classified as Category B2.

7.4  Assay

     7.4.1  Biobarrier preparation

            7.4.1.1  Biobarrier matrix preparation  -  must be completed  at
     least two hours prior to running assay.

            7.4.1.2  Place scintillation vial containing  biobarrier matrix
     powder on the  hot  plate pad.   Begin  spinning the stir bar before
     adding the diluent.

            7.4.1.3  Add the entire  contents of the biobarrier  diluent
     vial  slowly and constantly to the vial  of biobarrier matrix  powder.
     Make sure that the  stir bar is turning while adding  the diluent. The
     stir bar should  be turning rapidly, but  not so fast that the solution
     foams.

            7.4.1.4  Turn the heat  on  low; monitor the temperature of the
     solution as  it is warming.  Gradually increase the heat as necessary
     to warm the solution to 68°C  (±1°C)   to  solubilize  the  matrix.  This
     may take approximately 20  minutes.  DO  NOT allow the temperature  to
     exceed 70°C.

            7.4.1.5  While the  solution is warming, remove the tray of  24
     membrane discs from the refrigerator.   Remove  the tray  lid.
                             1120 -  5                       Revision  0
                                                            January 1995

-------
       7.4.1.6  After the biobarrier matrix solution reaches 68°C and
has been completely solubilized,  turn  off the heat  and move the vial
toward the edge  of the heat pad to keep it warm while aliquotting
into the membrane discs.

       7.4.1.7  Dispense  solubilized  solution  into membrane  discs.
Using the repeat pipettor, set to dispense  200  /xL.  Slowly fill the
pipette tip with  biobarrier solution, avoiding air bubbles.  Dispense
one aliquot back  into the biobarrier vial to ensure  proper subsequent
volume  delivery.    Wipe  the  tip before  dispensing each  aliquot.
Dispense 200  fj.1 into each disc,  ensuring that the entire membrane is
covered and no air bubbles have formed.  Any air bubbles in the gel
will alter the results of the  test,  therefore the disc  cannot be
used.

       7.4.1.8  Label the lid with the date, time, lot number, and
initials of the technician preparing the biobarrier. Wrap the filled
tray evenly  with plastic wrap;  do  not bunch  up  the  plastic  wrap
underneath the plate.

       7.4.1.9  Store the tray  at 2  - 8°C  for at  least  two  hours
before beginning any testing.  The biobarrier is stable for fourteen
days if wrapped and stored at 2 - 8°C.

7.4.2  Running the assay.

       7.4.2.1  On the  data  sheet,  complete the  lot  number,  date,
time, name of  technician, name  of chemical, whether it is solid or
liquid, and  pH of a  10% solution  diluted  in  water or appropriate
solvent.

       7.4.2.2  When ready to test,  remove one  tray of seven black-
capped scintillation vials from the kit box.

       7.4.2.3  Using a pipet-aid, dispense 22.0 mL of the Chemical
Detection System (CDS) into  each  of the six (6)  scintillation vials.
Make sure  that the CDS  is  at  room  temperature  (17  -  25°C)  before
using.  Dispense  12.0 ml  of  the  CDS  into the  seventh scintillation
vial and cap it.  This vial  will serve as a color control.

       7.4.2.4  Remove  the   tray of  24 membrane  discs  from  the
refrigerator.  Place on a tray of ice.

       7.4.2.5  Place disc  into  first scintillation vial.   Do not
allow the discs  to be in contact with the CDS  for longer than two
minutes before applying the test  sample.  Within two minutes, add 500
pi  (using the Lab Industries pipettor, or  equivalent) or  500 mg
(using spatula or tweezers)  of test  sample to disc.  Start timer the
instant the sample is added.

       Note:    Do not cap  the vials while test is in progress due to
       potential pressure build-up.
                        1120 - 6                       Revision 0
                                                       January 1995

-------
                  7.4.2.6  Watch the vial for three minutes, ensuring that the
           color reaction  is not missed  if  it  is  a Group I sample. Changes in
           the CDS may include various color changes, flaking or precipitation.

                  7.4.2.7  Add three more discs and samples to vials, staggering
           each  start  time  so  that  the  most  accurate  reaction times  are
           recorded.

                  7.4.2.8  Allow assay to run until color or physical reaction
           occurs.  Category A., and ft, samples  should be checked for reactions
           at 3 minutes, 1 hour and 4 hours.  Category A2 and B2 samples should
           be checked for  reactions at 3, 30 and 45 minutes.

                  7.4.2.9  At the first indication of the presence of a chemical
           reaction in  the CDS,  there  will  be  a color change produced beneath
           the bottom-center of each biobarrier disc. As soon as a reaction is
           observed, immediately record net time of each vial on the data sheet.

                  7.4.2.10 Run positive and negative controls in the other two
           vials, if desired, for GLP purposes.

           7.4.3  Assignment of Groups.

                  7.4.3.1  Category A,  and  B., samples are assigned to Group  I if
           a color change  is observed between zero and three minutes,  to Group
           II if a color change  is  observed  after  three minutes  and  up to one
           hour, and to Group III if a color change is observed after one hour
           and up to four  hours.  If no color change occurs  in four hours, the
           chemical is classified as Noncorrosive.

                  7.4.3.2  Category A2 and B2 samples are assigned to Group  I if
           a color change  is observed between zero and three minutes,  to Group
           II if a color  change is observed after three minutes and  up  to 30
           minutes, and to Group III  if a  color  change is  observed  after 30
           minutes and  up to  45 minutes.    If no  color change  occurs  in 45
           minutes, the chemical is classified  as Noncorrosive.
8.0   QUALITY CONTROL

      8.1  Samples should be analyzed in quadruplicate.  The test may be analyzed
in duplicate if a simple screening of corrosives and noncorrosives is all  that
is required.  However, it is recommended that for greater accuracy,  samples be
analyzed in quadruplicate.  It  is  suggested that positive and negative controls
be analyzed to conform with GLP.


9.0   METHOD PERFORMANCE

      9.1  Interlaboratory and  intralaboratory  studies were conducted with five
different laboratories.  Ten different chemicals were tested with  six replicates.
The data are presented in Table 1.
                                   1120 - 7                       Revision 0
                                                                  January 1995

-------
      9.2  More than 200 data points have been collected at  InVitro International
for six reference samples.  Statistical analysis of this data shows the standard
deviation for ethylenediamine and ferric chloride is about 5% of their respective
assay  times,  and  about   10%  for  maleic  anhydride,   sodium hydroxide,  and
dicyclohexylamine.  The standard deviation for sulfuric acid approaches 18% of
its assay time,  but  when taken into account that the mean assay  time  is less than
1 minute, a standard deviation of 0.13 minutes is actually a reflection of the
difficulty of measuring such brief timeperiods.

      9.3  The Corrositex assay has been used by more than 300 laboratories to
test approximately  4,000  test  materials  in its  first  phase of  utilization in
industry.   Diverse  chemicals  and formulations which  include  liquids,  solids,
insolubles and immiscibles have been studied from many major industries including
petrochemical, agrochemical, surfactant, textile, paper and pulp, electroplating
and water treatment.   Examples  of dermal  corrosion  values compaired  to pH for
selected compounds  are shown in Table 2.

      9.4  Data results from 1,050 samples  that have been tested using the assay
system were complied and  compared with in  vivo data.  Ninety-two percent of the
samples (965 samples) passed the Pre-qualification Test and were then analyzed
in the screening  test  and  the  assay.   Assay  was found to  be  highly concordant
with corrosive/noncorrosive in vivo results.   Of 406 corrosive samples with in
vivo data, 377  (93%) were  correctly  identified  as  corrosive by  assay.   Of 296
noncorrosive samples with  in vivo  data, 83%  were identified  as  noncorrosives,
demonstrating  the  ability of  this  in  vitro  method  to correctly  identify
corrosives and noncorrosives.  Assay was also shown to accurately predict Packing
Groups.  Six hundred out  of 702 samples (85.5%) were placed in the same Packing
Group as that  indicated  by in  vivo testing.   Only 38  test samples  out of 702
samples that had  in vivo data were found to underestimate  (5.4%).   Of these 38
samples, 28 were distinct samples and the  remainder were samples that had been
tested in more than one laboratory.  When taking  this information into account,
the percent of underestimation decreased to about 4%.


10.0  REFERENCES

1.    Code of Federal Regulations,  Transportation Title 49, Hazardous Materials
Table,  Section 172.101 (1991).

2.    Code of Federal  Regulations,  Transportation Title 49, Method  of Testing
Corrosion to the Skin, Part 173, Appendix A (1991).

3.    Schlesselman, J.J.  (1973)  Planning  a Longitudinal Study:  I.  Sample Size
Determination. J. Chron.  Dis.  26, 553-560.

4.    ASTM Standards on Precision and Bias for Various Applications,  "Standard
Practice for Conducting an Interlaboratory Study to Determine  the  Precision of
a Test Method"; ASTM: Philadelphia, PA, 1992; E 691-92.

5.    Gordon, V.C., Marvel 1,  J., and Maibach, H.  (1994).  Dermal  Corrosion, The
Corrositex  System,   A  DOT Accepted  Method  to  Predict  Corrosivity  of  Test
Materials.  In Vitro Toxicology. Ed. Mary Ann Liebert, 1994.
                                   1120 - 8                       Revision 0
                                                                  January 1995

-------
                                  TABLE  1
                               LABORATORY DATA
Dichloroacetyl chloride
    Laboratory
  Mean  of 6  Test
Results  (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
3.30
2.99
3.88
2.50
3.30
0.59
0.30
0.47
0.26
0.28
17.88%
10.03%
12.11%
10.40%
8.48%
Formic acid
    Laboratory
  Mean  of 6  Test
Results  (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
6.32
5.08
5.18
4.82
4.02
0.51
0.46
0.77
0.29
0.29
8.07%
9.06%
14.86%
6.02%
7.21%
Dichloroacetic acid
    Laboratory
  Mean  of  6 Test
Results (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
6.92
5.21
6.32
5.78
5.65
0.32
0.25
0.98
0.26
0.46
4.62%
4.80%
15.51%
4 . 50%
8.14%
Chloroacetic acid
    Laboratory
 Mean  of  6 Test
Results (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
5.46
7.83
4.95
6.91
4.95
0.36
0.00
0.35
0.94
0.34
6.59%
0 . 00%
7.07%
13.60%
6.87%
                                  1120 -  9
                                             Revision 0
                                             January 1995

-------
                             TABLE  1  (continued)
Dodecyltri chlorosllane

                     Mean of 6 Test
    Laboratory      Results  (minutes)
                        Standard
                       Deviation
                Relative  Standard
                    Deviation
1
2
3
4
5
10.78
11.59
11.22
11.96
10.98
0.05
0.36
0.82
0.56
0.29
0.46%
3.11%
7.31%
4.68%
2.64%
Ammonium hydrogen sulfate
    Laboratory
 Mean of 6 Test
Results  (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
10.47
9.02
13.82
11.17
7.88
0.56
0.33
1.0
0.93
0.26
5.35%
3.66%
7.24%
8.33%
3.30%
Ethylenediamine
    Laboratory
 Mean of  6 Test
Results  (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
17.24
21.33
26.12
20.76
22.48
0.38
0.53
1.30
0.19
1.40
2.20%
2.48%
4.98%
0.92%
6.23%
Aluminum chloride
    Laboratory
 Mean of 6 Test
Results  (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
11.91
21.33
26.12
20.76
22.48
0.41
0.53
1.30
0.19
1.40
3.44%
2 . 48%
4 . 98%
0.92%
6.23%
                                  1120  -  10
                                             Revision 0
                                             January 1995

-------
                            TABLE  1  (continued)
Acetic acid
    Laboratory
 Mean  of  6 Test
Results (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
28.52
27.00
34.98
36.30
29.67
0.87
0.00
0.22
0.41
0.62
3.05%
0.00%
0.63%
1.13%
2.09%
Dicyclohexylamine
    Laboratory
 Mean  of  6 Test
Results (minutes)
 Standard
Deviation
Relative Standard
    Deviation
1
2
3
4
5
181.73
168.83
210.70
159.04
126.75
0.47
9.11
7.68
7.58
0.62
0.26%
5.40%
3.64%
4.77%
0.49%
                                 1120 - 11
                                             Revision 0
                                             January 1995

-------
TABLE 2.  EXAMPLES OF DERMAL CORROSION VALUES FOR SELECTED COMPOUNDS
Compound Name
Acetic acid
Aluminum chloride
Ammonium hydroxide
Bromoacetic acid
Butyl ami ne
Citric acid
1,2-Diaminopropane
Dichloroacetic acid
Dicyclohexylamine
Di ethyl ami ne
Diethylene triamine
Ethanolamine
Ferric chloride
Formic acid
Hydrochloric acid
Hexanoic acid
Maleic acid
Mercaptoacetic acid
Nitric acid
Phosphoric acid
Potassium hydroxide
Propionic acid
Sodium hydroxide
Sodium metasil icate
Sulfuric acid
Thiophosphoryl chloride
Tri butyl ami ne
Trichlorotoluene
Triethanolamine
Triphosphoryl chloride
Concentration
(weight %)
99+
pure
10.00
55.60
40.00
20.00
99+
3.10
99.00
98.00
99.00
99+
98.00
33.90
35.00
99.00
99.00
15.10
90.00
85.00
pellets
99+
pellets
20.00
15.00
98.00
99+
99.00
60.00
98.00
PH1
0.00
2.92
12.37
0.93
12.96
1.28
12.06
0.98
9.57
13.86
12.01
11.82
3.00
0.62
0.00
3.00
1.30
1.60
0.00
0.00
14.00
0.35
13.81
13.28
0.00
5.81
10.70
3.32
11.02
5.80
Time
(minutes)
29.31
16.50
5.41
9.17
>240
47.65
21.67
37.63
210.00
5.89
34.00
21.68
21.30
>240
5.80
149.00
15.55
42.09
0.57
15.00
6.82
34.59
14.67
17.17
11.48
10.13
>240
>240
41.03
10.25
1  pH of a 10% solution of the compound  in  water.
                              1120  -  12
Revision 0
January 1995

-------
                              METHOD  i.^O
                            DERMAL CORROSION
                          7.2.1  Add sample to
                                 CDS.
                                  Is
                                there a
                            color change in
                              the reaction
                                vessel.
  Proceed wiTh
     assay.
       Is it
      a liquid
       or a
      solid?
                          7.3.2.1 Add sample
                           to tube and read
                             color change.
                             7.3.2.2 Assign
                               catergories.
7.3.1.1 Add sample
 to tube  and read
   color change.
                          7.3.2.3 Confirmation
                                  Test.
 7.3.1.2 Assign
   categories.
7.3.1.3 Confirmation
       test.
                           7.4.1.4 Heat assay
                            7.4.1.5 - 7.4.1.7
                             Apply solution
                              to membrane.
   Use other method
to determine corrosivity.
                               1120 -  13
          7.4.1.8 - 7.4.2.6 Add
           one membrane and
            apply to vial and
            look for change.
          7A.2.1 - 7.4.2.9 Add
            3 more disks and
           samples to vial and
         check for color reactions
           recording exact time
            of color change.
                                                             7.4.2.10 Run positive
                                                             and negative controls.
                                                              7.4.3 Assignment of
                                                                   Groups.
                                                                Revision 0
                                                                January 1995

-------
                                 METHOD 9050A

                             SPECIFIC CONDUCTANCE
1.0       SCOPE AND APPLICATION

          1.1   Method 9050  is used  to measure  the specific  conductance of
drinking, ground, surface,  and  saline waters and domestic and  industrial aqueous
wastes.  Method 9050  is not applicable to solid samples or to organic samples.


2.0       SUMMARY OF METHOD

          2.1   The specific conductance of a sample is measured using a self-
contained conductivity meter (Wheatstone bridge-type or equivalent).

          2.2   Whenever  possible,  samples are analyzed at 25°C.   If  samples are
analyzed at  different temperatures,  temperature corrections must  be made and
results reported at 25°C.


3.0       INTERFERENCES

          3.1   Platinum  electrodes can degrade and cause erratic results.  When
this happens,  as  evidenced by  erratic results or  flaking  off  of the platinum
black, the electrode  should be replatinized.

          3.2   The specific conductance cell  can  become coated with  oil  and
other materials.   It  is  essential  that the cell be  thoroughly  rinsed  and, if
necessary, cleaned between samples.


4.0       APPARATUS AND MATERIALS

          4.1   Self-contained   conductivity   instruments:      an   instrument
consisting  of  a  source  of  alternating current,  a  Wheatstone bridge,  null
indicator, and  a conductivity  cell or  other instrument  measuring the ratio of
alternating current through the cell  to voltage across it.  The latter has the
advantage of a linear reading of conductivity.  Choose an instrument capable of
measuring conductivity with an  error not exceeding 1% or 1 uS/cm, whichever is
greater.

          4.2   PI at i num-el ectrode or non-platinum-electrode specific conductance
cell.

          4.3   Water  bath.

          4.4   Thermometer:   capable of being  read to the  nearest 0.1°C  and
covering the  range 23"C to  27°C.   An  electrical  thermometer  having  a  small
thermistor sensing element is convenient because of its rapid response.

                                   9050A -  1                       Revision 1
                                                                  January 1995

-------
5.0       REAGENTS

          5.1    Reagent  grade chemicals  shall  be used  in  all tests.   Unless
otherwise  indicated,  it is  intended  that all  reagents  shall conform  to the
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where  such  specifications are available.  Other grades  may be used,
provided it is first  ascertained that the reagent is  of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

          5.2    Conductivity water:  Pass distilled water through a mixed-bed
deionizer and discard first  1,000  ml.  Conductivity should  be  less  than 1 uS/cm.

          5.3    Standard potassium  chloride   (0.0100  M):    Dissolve  0.7456  g
anhydrous KC1  in conductivity water and  make up to  1,000  ml at 25°C.   This
solution will have a specific conductance of 1,413 uS/cm at  25°C.


6.0       SAMPLE COLLECTION, PRESERVATION, AND HANDLING

          6.1    All   samples must  be   collected  using  a  sampling   plan  that
addresses the considerations discussed in Chapter Nine of this manual.

          6.2    All  sample  containers  must be prewashed and  thoroughly rinsed.
Both plastic and glass containers are suitable.

          6.3    Aqueous  samples should  be stored at  4°C and  analyzed within 28
days.


7.0       PROCEDURE

          7.1    Determination of cell constant:  Rinse  conductivity cell with at
least three portions  of 0.01 N KC1 solution.   Adjust  temperature of a fourth
portion  to  25.0 +  0.1°C.    Measure  resistance  of  this   portion   and  note
temperature.  Compute cell   constant, C:


                C =  (0.001413)(RKC1)  1  +  0.0191 (t - 25)

                where:

                             = measured resistance,  ohms;  and

                        t =  observed temperature, °C.


          7.2    Conductivity measurement:  Rinse  cell with one or more portions
of  sample.   Adjust  temperature of  a  final  portion  to 25.0  + 0.1°C.  Measure
sample resistance or conductivity and note temperature.


                                   9050A  - 2                       Revision 1
                                                                  January 1995
4

-------
          7.3   Calculation:  The temperature coefficient of most waters  is only
approximately the same as that of standard KC1 solution; the more the temperature
of measurement deviates from  25.0°C, the greater the uncertainty in applying the
temperature correction.  Report all conductivities at 25.0°C.

                7.3.1    When sample resistance is measured, conductivity  at 25"C
i s:
               K =
                        (1,000,000)(C)
                   Rm1 + 0.0191 (t - 25)
                where:

                      K = conductivity,  uS/cm;
                      C = cell  constant,  cm-L;
                      Rm = measured resistance  of sample,  ohms;  and
                      t = temperature of  measurement.


                7.3.2   When sample  conductivity  is  measured,  conductivity at
    25°C  is:


                   (K  )(1,000,000)(C)
               K =   m
                   I + 0.0191 CT^ Zb)

                      where:

                       Km =        measured conductivity,  uS at t°C, and other
                                   units are defined as above.

          NOTE:  If conductivity readout is in uS/cm,  delete the factor 1,000,000
          in the numerator.

8.0  QUALITY CONTROL

          8.1    Refer to Chapter One  for specific  quality control  procedures.

          8.2    Analyze  an  independently prepared  check  standard  to  verify
calibration.

          8.3    Analyze one duplicate  sample  for  every  10  samples.
                                   9050A  -  3                       Revision 1
                                                                  January 1995

-------
9.0  METHOD PERFORMANCE
          9.1    Three  synthetic samples were tested with the following results:

Conduc-
tivity
uS/cm
147.0
303.0
228.0


No. of
Results
117
120
120
Relative
Standard
Deviation
%
8.6
7.8
8.4

Relative
Error
%
9.4
1.9
3.0
10.0  REFERENCES

1.   Standard Methods for  the Examination of  Water  and Wastewater,  16th  ed.
(1985), Method 205.
                                   9050A -  4                       Revision 1
                                                                  January 1995

-------
    METHOD 9050A

SPECIFIC CONDUCTANCE
7.1 Determination of
cell constant.
>
r
7.2 Measure
conductivity;
note temperature.
>
r
7.3 Perform
calculations.
3
r
      9050A - 5
Revision 1
January 1995

-------
                                 METHOD 9095A

                           PAINT FILTER LIQUIDS  TEST
1.0  SCOPE AND APPLICATION

      1.1   This method is used to determine the presence of free liquids in a
representative sample of waste.

      1.2   The method is used to determine compliance with 40 CFR 264.314 and
265.314.

2.0  SUMMARY OF METHOD

      2.1   A predetermined amount of material  is  placed  in  a paint filter.  If
any portion of the material  passes through  and  drops from the filter within the
5-min test period, the material is deemed to contain free liquids.

3.0  INTERFERENCES

      3.1   Filter media  were observed  to separate from  the  filter  cone  on
exposure to alkaline materials.  This development causes no problem if the sample
is not disturbed.

      3.2   Temperature can  affect  the  test results if the test is  performed
below the freezing point of any liquid in the sample.   Tests must be performed
above the freezing point and can,  but is not required to, exceed room temperature
of 25° C.

4.0  APPARATUS AND MATERIALS

      4.1   Conical  paint filter:   Mesh  number 60 +/- 5%  (fine  meshed size).
Available at local paint stores such as Sherwin-Williams and Glidden.

      4.2   Glass funnel:   If the paint filter, with the waste,  cannot sustain
its weight on the ring stand,  then a fluted  glass funnel or glass funnel with a
mouth large enough to  allow at least 1 in. of the filter mesh to protrude should
be used to support the filter.  The  funnel  is to be fluted or have a large open
mouth in order to support  the paint  filter yet  not interfere with the movement,
to the graduated cylinder, of the liquid that passes through the filter mesh.

      4.3   Ring stand and ring,  or tripod.

      4.4   Graduated cylinder or beaker:  100-mL.

5.0   REAGENTS

      5.1   None.
                                  9095A  -  1                       Revision 1
                                                                  January 1995

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   All samples must be collected according to the directions in Chapter
Nine of this manual.

      6.2   A  100-mL or 100-g representative sample  is  required  for the test.
If it is not possible to obtain a sample of 100  mL  or  100 g that is sufficiently
representative of the waste, the analyst may use larger size samples in multiples
of 100 mL or 100 g, i.e.,  200, 300, 400 mL or g.   However,  when larger samples
are used,  analysts  shall divide the sample  into 100-mL or 100-g portions and test
each portion separately.  If any  portion contains free liquids, the entire sample
is considered to have free liquids.  If the sample is measured  volumetrically,
then it should lack major air spaces or voids.

7.0  PROCEDURE

      7.1   Assemble test apparatus as shown in Figure 1.

      7.2   Place sample in the  filter.  A funnel may be  used to provide support
for the  paint  filter.   If  the  sample is of such light bulk  density  that it
overflow  the  filter,  then the sides  of  the  filter can be  extended upward by
taping filter paper to  the inside of the filter and above the  mesh.  Settling the
sample into the paint filter may be facilitated by lightly tapping the side of
the filter as  it is being filled.

      7.3   In  order  to  assure  uniformity and standardization  of  the test,
material such as sorbent pads or pillows which do not conform to  the shape of the
paint filter, should be cut into small pieces and poured into the filter.  Sample
size reduction may be accomplished by cutting the sorbent material with scissors,
shears, knife,  or  other such  device  so as to preserve as  much  of the original
integrity of  the sorbent  fabric  as  possible.   Sorbents enclosed in  a fabric
should be mixed  with the  resultant fabric pieces.   The particles to be tested
should be reduced smaller than 1 cm (i.e., should  be capable of passing through
a  9.5  mm  (0.375 inch)  standard sieve).   Grinding  sorbent  materials should be
avoided as this may destroy the  integrity  of the sorbent and produce many "fine
particles" which would normally not be present.

      7.4   For  brittle materials  larger  than  1 cm that do  not conform to the
filter, light  crushing to reduce oversize particles is acceptable if it is not
practical to  cut the  material.   Materials such as  clay,  silica  gel,  and some
polymers may fall  into this category.

      7.5   Allow  sample to drain  for  5 min into the graduated cylinder.

      7.6    If any portion  of  the  test  material  collects in  the graduated
cylinder in the 5-min period, then the material  is deemed to contain free liquids
for purposes of  40  CFR 264.314  and 265.314.

8.0  QUALITY CONTROL

      8.1   Duplicate  samples should  be analyzed on a routine  basis.
                                   9095A -  2                      Revision 1
                                                                  January 1995

-------
9.0  METHOD PERFORMANCE

      9.1   No data provided.

10.0  REFERENCES

      10.1  None provided.
                                   9095A -  3                       Revision 1
                                                                  January 1995

-------
RING STAND —
                                      ^-FUNNEL
                                          /PAINT FILTER
                                 i— GRADUATED CYLINDER
          Figure 1.  Paint  filter test apparatus.
                      9095A - 4
Revision 1
January 1995

-------
       METHOD 9095A
PAINT FILTER  LIQUIDS TEST
     Start
1
r
7.1 Assemble
test apparatus.
^
r
7.2 Place sample
in filter.
i
r
7.3 Allow sample
to drain into
graduated cylinder.
    7.4 Did
    any test
material collect
  in graduated
   cylinder?
7.4 Material is
deemed to contain
free liquids; see 40
CFR 264.314 or
265.314.
i
r
     Stop
        9095A - 5
Revision  1
January  1995

-------
                                CHAPTER EIGHT

                   METHODS  FOR  DETERMINING CHARACTERISTICS
      Methods for determining the characterisitics of Ignitability for liquids,
Corrosivity for liquids, and  Toxicity are  included.   Guidance for determining
Toxic Gas Generation is found in Chapter Seven,  Sections 7.3.3 and 7.3.4.
                                   EIGHT  -  1                       Revision 2
                                                                  January 1995

-------
8.1   Ignitability

      The following methods are found in Section 8.1:
            Method 1010:      Pensky-Martens Closed-Cup Method for Determining
                              Ignitability
            Method 1020A:     Setaflash  Closed-Cup   Method  for  Determining
                              Ignitability
                                   EIGHT  -  2                       Revision 2
                                                                  January 1995

-------
8.2   Corrosivity

      The following methods are found in Section 8.2:

            Method 9040B:     pH Electrometric Measurement
            Method 1110:      Corrosivity Toward Steel
                                   EIGHT  -  3                       Revision 2
                                                                  Januarv 1995

-------
8.3   Reactivity

      Refer to guidance  given  in Chapter Seven, especially  Section  7.3.3 and
7.3.4.
                                   EIGHT  - 4                       Revision 2
                                                                  January 1995

-------
8.4   Toxicity

      The following methods are found in Section 8.4:
            Method 1310A:     Extraction  Procedure  (EP) Toxicity  Test Method
                              and Structural Integrity Test
            Method 1311:      Toxicity Characteristic Leaching Procedure
                                   EIGHT -  5                      Revision 2
                                                                  January 1995

-------
                            CHAPTER TEN

                         SAMPLING METHODS
The following methods are found in Chapter Ten:
      Method 0010:
            Appendix A:
            Appendix B:

      Method 0011:

      Method 0020:
      Method 0023A:
      Method 0030:
      Method 0031:

      Method 0040:
      Method 0050:
      Method 0051:
      Method 0100:
Modified Method 5 Sampling Train
      Preparation of XAD-2 Sorbent Resin
      Total  Chromatographable Organic  Material
      Analysis
Sampling   for   Formaldehyde   Emissions   from
Stationary Sources
Source Assessment Sampling System (SASS)
Sampling  Method  for  Polychlorinated  Dibenzo-p-
Dioxins    and   Polychlorinated    Dibenzofuran
Emissions from Stationary Sources
Volatile Organic Sampling Train
Sampling  Method  for Volatile  Organic Compounds
(SMVOC)
Sampling   of    Principal    Organic   Hazardous
Constituents   from   Combustion   Sources   Using
Tedlar® Bags
Isokinetic HC1/C12 Emission Sampling Train
Midget Impinger HC1/C12 Emission  Sampling Train
Sampling  for  Formaldehyde  and  Other  Carbonyl
Compounds in Indoor Air
                             SIX  -  1
                                    Revision 2
                                    January 1995

-------
                                  METHOD  0011

          SAMPLING  FOR  FORMALDEHYDE  EMISSIONS  FROM  STATIONARY  SOURCES
1.0  SCOPE AND APPLICATION

      1.1  This  method  is applicable to the determination  of Destruction and
Removal  Efficiency  (ORE) of  formaldehyde,  CAS  Registry number  50-00-0,  and
possibly other aldehydes  and ketones from stationary sources  as specified in the
regulations.    The methodology has been applied  specifically  to formaldehyde.
However, many laboratories have extended the application  to other aldehydes and
ketones.

      1.2  Compounds derivatized with 2,4-dinitrophenylhydrazine can be detected
as low as  6.4  x  10"8  Ibs/cu ft (1.8 ppbv)  in stack  gas  over  a 1 hour sampling
period, sampling approximately 45 cu ft.

      1.3  This method is restricted to use by, or under the close supervision
of, analysts  experienced  in sampling organic compounds in air.  Each analyst must
demonstrate the ability to generate acceptable results with this method.


2.0  SUMMARY OF METHOD

      2.1  Gaseous and particulate pollutants are withdrawn isokinetically from
an  emission  source  and   are  collected in  aqueous acidic  2,4-dinitrophenyl-
hydrazine.  Formaldehyde present in the emissions reacts with the 2,4-dinitro-
phenylhydrazine to form the formaldehyde dinitrophenylhydrazone derivative.  The
dinitrophenylhydrazone derivative is extracted,  solvent-exchanged, concentrated,
and then analyzed by high performance liquid chromatography (HPLC).


3.0  INTERFERENCES

      3.1  A decomposition product of 2,4-dinitrophenylhydrazine, 2,4-dinitro-
aniline, can  be  an  analytical  interferant  if  concentrations are high.   2,4-
Dinitroaniline can coelute with the 2,4-dinitrophenylhydrazone of formaldehyde
under high performance liquid chromatography conditions used for the analysis.
High concentrations of highly oxygenated compounds, especially acetone, that have
the  same  retention   time  or   nearly  the   same  retention   time   as   the
dinitrophenylhydrazone of  formaldehyde, and  that also absorb  at  360  nm,  will
interfere with the analysis.

      3.2  Formaldehyde,  acetone,  and  2,4-dinitroaniline contamination of the
aqueous  acidic  2,4-dinitrophenylhydrazine   (DNPH)  reagent   is   frequently
encountered.   The reagent must be prepared within five days of use in the field
and must  be  stored  in   an uncontaminated  environment  both  before and  after
sampling,  in order to minimize  blank problems.   Some  concentration  of acetone
contamination  is unavoidable,  because  acetone  is  ubiquitous  in laboratory and
field operations.  However, the acetone contamination  must be minimized.
                                   0011 - 1                         Revision 0
                                                                  January 1995

-------
4.0  APPARATUS AND MATERIALS

      4.1   This  sampling  train configuration  is  adapted  from  EPA Method  5
procedures.  The  sampling  train consists of the following  components:   Probe
nozzle, pitot tube, differential pressure gauge,  metering system, barometer, and
gas density determination equipment.  A schematic of the sampling train is shown
in Figure 1.

            4.1.1  Probe Nozzle  - The probe nozzle shall be quartz or glass with
      sharp, tapered  (30°  angle)  leading edge.   The taper  shall  be  on  the
      outside  to  preserve  a  constant inner  diameter.   The  nozzle shall  be
      buttonhook  or  elbow  design.   A  range  of  nozzle sizes  suitable  for
      isokinetic  sampling should be  available in increments of 0.16 cm (1/16
      in.), e.g., 0.32 to 1.27 cm (1/8 to 1/2 in.), or larger if higher volume
      sampling trains are used.  Each  nozzle  shall  be calibrated  according to
      the procedures outlined in Sec. 8.1.

            4.1.2  Probe Liner - Borosilicate glass or quartz shall be used for
      the probe liner.   The tester should not allow  the temperature in the probe
      to exceed 120 ±14'C (248 ±25'F).

            4.1.3  Pitot Tube - The pitot  tube shall  be Type S  or  any other
      appropriate device.  The Type S pitot tube shall be made of metal  tubing
      (e.g., stainless  steel).   It  is  recommended  that  the  external  tubing
      diameter be between 0.48  and 0.95  cm.   There  shall  be an equal  distance
      from the base of  each  leg to its  face-opening  plane;  it is recommended
      that this  distance be between  1.05  and  1.50 times the  external  tubing
      diameter.   The  face openings  of  the  pitot  tube shall, preferably,  be
      aligned  but  slight misalignments of the  openings are  permissible.   The
      Type S pitot tube assembly shall have a known coefficient,  determined as
      outlined in Sec. 4 of EPA Method 2.  The pitot tube shall be attached to
      the probe to allow constant  monitoring of the stack  gas velocity.   The
      impact (high pressure)  opening  plane of the pitot tube shall be even with
      or above the nozzle  entry plane (see EPA Method 2, Figure  2-6b)  during
      sampling.

            4.1.4  Differential  Pressure  Gauge -  The differential pressure gauge
      shall be an inclined manometer or equivalent device as described  in Sec.
      2.2  of EPA Method 2.   One  manometer shall  be used  for  velocity-head
      readings and the other for orifice differential pressure readings.

            4.1.5  Impingers  -  The sampling train requires  a  minimum  of four
      impingers,  connected  as   shown  in  Figure  1,  with  ground  glass  (or
      equivalent)  vacuum-tight  fittings.   For  the  first,  third, and  fourth
      impingers, use the Greenburg-Smith design, modified by replacing  the tip
      with a 1.3  cm  inside  diameter  (1/2 in.)  glass tube extending  to  1.3 cm
      (1/2 in.) from the bottom of the flask.   For the  second impinger, use a
      Greenburg-Smith  impinger  with  the standard  tip.   Place a  thermometer
      capable of measuring  temperature to within 1°C  (2°F) at the outlet of the
      fourth impinger for monitoring purposes.


                                   0011  - 2                         Revision 0
                                                                  January 1995

-------
      4.1.6  Metering System  -  The  necessary  components of the metering
system  are  a  vacuum  gauge,  leak-free  pump,  thermometers  capable  of
measuring  temperature  within  3°C  (5.4°F),  dry-gas  meter  capable  of
measuring volume to within  1%, and related equipment as  shown  in Figure 1.
At a minimum, the pump should be capable of 4 cfm free flow,  and the dry
gas  meter  should  have  a  recording capacity  of  0-999.9  cu ft with  a
resolution of 0.005 cu ft.  Other metering systems may be used which are
capable of maintaining sample volumes to within 2%.  The metering system
may  be used  in  conjunction  with  a  pitot  tube  to  enable checks  of
isokinetic sampling rates.

      4.1.7  Barometer - The barometer may be mercury,  aneroid, or other
barometer capable of  measuring  atmospheric  pressure  to within  2.5  mm Hg
(0.1 in. Hg).  In many cases,  the barometric reading may be obtained from
a nearby National  Weather Service Station, in which case the station value
(which is the absolute barometric pressure) is  requested and an adjustment
for elevation differences between the weather station and sampling point
is applied at a  rate  of minus 2.5  mm Hg  (0.1  in.  Hg)  per  30 m (100 ft)
elevation increase  (vice versa for elevation decrease).

      4.1.8   Gas  Density  Determination Equipment  -   The  gas  density
determination equipment includes a temperature sensor and pressure gauge
(as described in Sections 2.3  and 2.4  of  EPA Method 2), and gas analyzer,
if  necessary (an  Orsat  of  Fyrite type  combustion  gas  analyzer,  or
equivalent.  For  analyzer maintenance and operation procedures, follow the
instructions recommended  by the manufacturer).   The  temperature sensor
ideally should be permanently  attached to the pitot tube or sampling probe
in a fixed configuration such that  the  tip  of the sensor extends beyond
the  leading  edge  of  the  probe sheath  and  does  not  touch  any  metal.
Alternatively,  the sensor may  be attached just prior to use in the field.
Note, however,  that if  the  temperature  sensor is  attached  in the field,
the sensor must be placed in an interference-free arrangement  with respect
to the Type S pitot tube openings  (see  EPA  Method 2,  Figure  2-7).   As a
second alternative, if  a difference of  no  more  than  1% in  the average
velocity measurement is to be introduced, the  temperature gauge need not
be attached to  the probe or pitot tube.

4.2  Sample Recovery

      4.2.1  Probe Liner  - Probe   nozzle  and brushes;  Teflon® bristle
brushes with stainless steel wire handles are  required.  The  probe brush
shall have extensions of stainless  steel, Teflon®,  or  inert  material  at
least as  long  as the  probe.   The  brushes  shall  be properly  sized and
shaped to brush  out the probe  liner, the  probe nozzle, and the impingers.

      4.2.2  Wash Bottles - Three wash bottles are required.   Teflon® or
glass wash bottles are recommended.  Polyethylene wash bottles should not
be used because  organic contaminants may be extracted by exposure to the
organic solvents used for sample recovery.
                             0011 - 3                         Revision 0
                                                            January 1995

-------
      4.2.3  Graduated Cylinder and/or Balance - A graduated cylinder or
balance is required to measure condensed water to the nearest  1 mL or 1 g.
Graduated  cylinders  shall  have  divisions  not  greater  than  2  mL.
Laboratory balances capable of weighing to ±0.5 g are required.

      4.2.4  Amber Glass Storage Containers - One-liter wide-mouth amber
flint glass bottles with Teflon®-!ined caps are required to store impinger
water samples.  The bottles must be sealed with Teflon® tape.

      4.2.5  Rubber Policeman and Funnel - A rubber policeman and funnel
are  required to  aid  in   the  transfer  of materials  into  and out  of
containers in the field.

4.3  Reagent Preparation

      4.3.1  Bottles/Caps  - Amber 1  -  4  L bottles with Teflon®-!ined caps
are required for  storing cleaned DNPH solution.   Additional  4-L bottles
are required to collect waste organic solvents.

      4.3.2  Large Glass Container  - At  least  one large glass (8 to 16 L)
is required for mixing the aqueous acidic DNPH solution.

      4.3.3  Stir  Plate/Large Stir  Bars/Stir  Bar Retriever - A magnetic
stir plate and large  stir bar are required for the mixing of the aqueous
acidic DNPH  solution.  A  stir bar retriever  is  needed for removing the
stir bar from the large container holding the DNPH solution.

      4.3.4   Buchner  Filter/Filter Flask/Filter Paper -  A large filter
flask (2-4 L) with  a  buchner filter,  appropriate rubber  stopper, filter
paper, and connecting  tubing  are required for filtering the aqueous acidic
DNPH solution prior to cleaning.

      4.3.5   Separatory Funnel   -  At  least one  large  separatory funnel
(2 L) is required for cleaning the DNPH prior to use.

      4.3.6  Beakers - Beakers (150 mL, 250 mL, and 400 mL) are useful for
holding/measuring organic  liquids  when  cleaning  the aqueous  acidic DNPH
solution and for weighing DNPH crystals.

      4.3.7  Funnels - At least one large funnel is needed  for pouring the
aqueous acidic DNPH into the separatory funnel.

      4.3.8  Graduated Cylinders - At least one large graduated cylinder
(1 to 2 L) is required for measuring organic-free reagent water and acid
when preparing the DNPH solution.

      4.3.9   Top-loading  Balance  - A one-place top loading balance is
needed for weighing  out the DNPH  crystals used to  prepare  the aqueous
acidic DNPH solution.
                             0011 - 4                         Revision 0
                                                            January 1995

-------
            4.3.10  Spatulas  -  Spatulas  are  needed  for weighing out DNPH when
      preparing the aqueous DNPH solution.

      4.4  Crushed Ice - Quantities of crushed  ice ranging from  10-50 Ib may be
necessary during  a  sampling  run,  depending  upon  ambient temperature.   Samples
which have been taken must be stored and shipped cold; sufficient ice for this
purpose must be allowed.


5.0  REAGENTS

      5.1  Reagent Grade Chemicals - Reagent grade chemicals shall be  used in all
tests.   Unless otherwise  indicated,  it  is  intended  that all  reagents  shall
conform to the  specifications of the Committee on  Analytical  Reagents  of the
American Chemical  Society, where such specifications  are available. Other grades
may be used, provided  it is first ascertained that the reagent  is  of sufficiently
high  purity  to  permit  its  use  without  lessening  the  accuracy  of  the
determination.

      5.2  Organic-free Reagent Water - All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3  Silica Gel  - Silica gel  shall  be indicating  type, 6-16 mesh.  If the
silica gel has  been used previously, dry at 175°C  (350°F) for  2 hours before
using.  New silica gel may be used as received.  Alternatively, other types of
desiccants (equivalent or better) may be used.

      5.4 2,4-Dinitrophenylhydrazine (DNPH),  [2,4-(02N)2C6H3]NHNH2 - The quantity
of water may vary from  10 to 30%.

            5.4.1   The  2,4-dinitrophenylhydrazine reagent must  be prepared in
      the laboratory within five days of sampling use in the field.  Preparation
      of DNPH can also be done in the field,  with consideration of appropriate
      procedures  required  for safe handling  of solvent in the  field.   When a
      container of prepared DNPH reagent is opened in the field,  the contents of
      the opened  container should  be  used  within  48  hours.    All  laboratory
      glassware must be washed with detergent and water and rinsed with water,
      methanol, and methylene chloride prior to use.

NOTE: DNPH crystals and  DNPH  solution  are potential  carcinogens and should be
      handled with plastic gloves at all times,  with  prompt and extensive use of
      running water in case of skin exposure.

            5.4.2   Preparation  of Aqueous Acidic DNPH  Derivatizing  Reagent -
      Each batch  of DNPH reagent should be prepared and  purified within  five
      days of sampling, according to the procedure described  below.

NOTE: Reagent bottles for storage of cleaned DNPH derivatizing solution must be
      rinsed with acetonitrile  and dried before use.   Baked glassware  is not
      essential for  preparation of DNPH  reagent.   The glassware must  not be


                                   0011 - 5                         Revision 0
                                                                  January 1995

-------
rinsed  with  acetone  or   an   unacceptable   concentration   of  acetone
contamination will  be  introduced.    If  field  preparation  of DNPH  is
performed, caution must be exercised in avoiding acetone contamination.

            5.4.2.1   Place  an  8 L  container  under  a  fume  hood  on  a
      magnetic stirrer.  Add a  large  stir  bar and fill the container half
      full of organic-free reagent water.  Save  the empty bottle from the
      organic-free reagent water.  Start the stirring bar and adjust the
      stir rate to be  as  fast  as possible.   Using  a graduated cylinder,
      measure 1.4 L of concentrated  hydrochloric acid.   Slowly pour the
      acid into the stirring water.   Fumes may be generated and the water
      may become warm.  Weigh the DNPH crystals to ±0.1  g  (see Table  1 for
      approximate amounts)  and  add to  the stirring  acid  solution.   Fill
      the 8 L container to  the  8 L  mark with organic-free  reagent water
      and stir  overnight.   If  all  of  the DNPH crystals have dissolved
      overnight,  add   additional  DNPH  and  stir   for  two  more  hours.
      Continue the process of adding DNPH with additional stirring until
      a saturated  solution  has  been formed.   Filter  the  DNPH solution
      using vacuum filtration.   Gravity filtration  may be  used,  but a much
      longer time is required.   Store  the filtered  solution  in an amber
      bottle at room temperature.

            5.4.2.2  Within five days of proposed  use,  place about 1.6 L
      of the DNPH reagent in a 2 L separatory funnel.  Add  approximately
      200 ml  of methylene  chloride  and  stopper the  funnel.    Wrap the
      stopper of  the  funnel with paper  towels to absorb  any leakage.
      Invert and vent the funnel.   Then  shake  vigorously for 3 minutes.
      Initially, the funnel  should  be  vented frequently (every  10  -  15
      seconds).    After  the layers  have separated,  discard  the  lower
      (organic) layer.

            5.4.2.3    Extract  the  DNPH  a  second  time  with  methylene
      chloride and finally with cyclohexane.  When the cyclohexane layer
      has separated from the DNPH reagent, the cyclohexane  layer will  be
      the top layer in  the separatory funnel.  Drain the lower layer (the
      cleaned extracted DNPH reagent solution) into an  amber bottle that
      has been rinsed with acetonitrile and allowed to dry.

      5.4.3  DNPH Reagent Check - Take  two aliquots of the extracted DNPH
reagent.  The size of  the  aliquots  is  dependent upon the exact sampling
procedure used,  but  100 ml  is  reasonably representative.    Analyze one
aliquot of  the  reagent according to the  procedure of Method  8315  as  a
Quality Control  check  to ensure  that  the background  in the  reagent  is
acceptable for field use.  Save the other aliquot  of aqueous acidic DNPH
for use as a method blank when the analysis is performed.

      5.4.4  Shipment  to  the Field  -  Tightly cap  the  bottle containing
extracted DNPH reagent using a  Teflon®-!ined  cap.   Seal  the bottle with
Teflon® tape.  After the bottle is labeled, the  bottle may be placed in a
friction-top can (paint can or  equivalent) containing a 1-2 inch layer of
granulated charcoal and stored at ambient temperature until  use.

                             0011 - 6                         Revision 0
                                                            January 1995

-------
                  5.4.4.1  If the DNPH  reagent  has  passed the Quality Control
            criteria, the  reagent may be packaged  to  meet  necessary shipping
            requirements and sent to the sampling  area.  If the Quality Control
            criteria are not met, the  reagent  solution may  be re-extracted or
            the  solution  may   be   re-prepared  and  the  extraction  sequence
            repeated.

                  5.4.4.2  If the DNPH reagent is not used in the field within
            five days of extraction,  an aliquot may be taken  and  analyzed as
            described in Method  8315.   If the reagent meets the Quality Control
            requirements, the reagent  may be used.  If the  reagent does not meet
            the Quality Control  requirements, the  reagent  must be discarded and
            new reagent must be prepared and tested.

            5.4.5   Calculation  of Acceptable Concentrations  of  Impurities in
      DNPH  Reagent  - The  acceptable  impurity  concentration  (AIC,  ng/ml)  is
      calculated from  the expected analyte  concentration in the  sampled gas
      (EAC, ppbv),  the volume of air that will  be sampled at standard conditions
      (SVOL, L), the formula weight of the analyte (FW, g/mol), and the volume
      of DNPH reagent that will  be used  in the impingers  (RVOL, ml):

      AIC = 0.1 x [EAC x SVOL x FW/24.4 x (FW + 180J/FW](RVOL/1,000)

      where:
            0.1 is the acceptable contaminant concentration,
            24.4 is a factor relating ppbv to g/L,
            180 is a factor relating underivatized to derivatized analyte
            1,000 is a unit conversion factor.

            5.4.6  Disposal of Excess  DNPH Reagent - Excess DNPH reagent may be
      returned to the laboratory and  recycled  or  treated  as  aqueous waste for
      disposal purposes.  2,4-Dinitrophenylhydrazine is a flammable solid when
      dry, so water should not  be evaporated from the solution of the reagent.

      5.5  Field Spike  Standard  Preparation  -  To  prepare  a  formaldehyde field
spiking standard at 4010 mg/L,   use  a  500 juL  syringe to transfer  0.5 ml of 37%
by weight  of formaldehyde (401  g/L)  to  a  50  mL volumetric  flask containing
approximately 40 mL of methanol.  Dilute to 50 mL with methanol.

      5.6     Hydrochloric  Acid,   HC1   -  Reagent  grade  hydrochloric  acid
(approximately 12N) is required  for acidifying the aqueous DNPH solution.

      5.7  Methylene Chloride, CH2C12 - Methylene chloride  (suitable for residue
and pesticide  analysis,  GC/MS,  HPLC,  GC, Spectrophotometry or equivalent)  is
required for cleaning the aqueous acidic DNPH solution, rinsing glassware, and
recovery of sample trains.

      5.8  Cyclohexane,  C6H12  - Cyclohexane (HPLC grade)  is  required for cleaning
the aqueous acidic DNPH solution.
                                   0011 - 7                         Revision 0
                                                                  January 1995

-------
NOTE: Do  not  use  spectroanalyzed  grades  of  cyclohexane   if  this  sampling
      methodology is extended to aldehydes  and  ketones with four or more carbon
      atoms.

      5.9 Methanol, CH3OH - Methanol (HPLC grade or equivalent) is necessary for
rinsing glassware.

      5.10   Acetonitrile,  CH3CN  -  Acetonitrile (HPLC grade  or  equivalent)  is
required for rinsing glassware.

      5.11   Formaldehyde,  HCHO  -  Formaldehyde (analytical   reagent  grade,  or
equivalent)   is required  for preparation  of standards.    If other  aldehydes  or
ketones are  used, analytical reagent grade, or equivalent,  is required.


6.0  SAMPLE  COLLECTION, PRESERVATION,  AND HANDLING

      6.1  Because of the complexity of this method,  field  personnel  should  be
trained in and experienced  with the  test procedures in order to obtain reliable
results.

      6.2  Laboratory Preparation

            6.2.1   All  the  components  shall  be maintained  and  calibrated
      according to  the  procedure described in APTD-0576 (Air Pollution Technical
      Document, see references), unless otherwise specified.

            6.2.2  Weigh  several  200 to 300 g portions of  silica gel in airtight
      containers to  the  nearest 0.5  g.    Record  on  each container  the  total
      weight  of the   silica  gel  plus  containers.    As  an  alternative  to
      preweighing  the  silica gel,  it may  instead be weighed directly  in  the
      impinger or sampling holder just prior to train assembly.

      6.3  Preliminary Field Determinations

            6.3.1  Select the sampling site and the minimum number of sampling
      points according to  EPA  Method  1 or  other relevant criteria.   Determine
      the stack  pressure,  temperature,  and range of velocity heads  using  EPA
      Method 2.  A  leak-check of the pitot lines according to  EPA Method 2, Sec.
      3.1, must  be performed.   Determine the stack gas  moisture content  using
      EPA Approximation Method 4 or its alternatives  to  establish estimates  of
      isokinetic sampling-rate settings.  Determine the stack gas dry molecular
      weight, as described in EPA Method 2,  Sec. 3.6.  If integrated EPA Method
      3 sampling is used  for molecular weight determination, the integrated bag
      sample shall  be taken  simultaneously with, and  for the same total length
      of time as, the sample run.

            6.3.2  Select a nozzle  size based on the  range of velocity heads so
      that it  is not necessary to  change the nozzle  size in  order to maintain
      isokinetic sampling  rates  below 28 L/min  (1.0  cfm).   During the run,  do


                                   0011 - 8                         Revision 0
                                                                  January 1995

-------
      not change the  nozzle.  Ensure that the  proper differential pressure gauge
      is chosen for the range of velocity heads encountered  (see Sec. 2.2 of EPA
      Method 2).

            6.3.3  Select a suitable probe liner  and  probe length  so  that all
      traverse points can be sampled.  For large stacks,  to  reduce the length of
      the probe, consider sampling from opposite sides of the stack.

            6.3.4  A minimum  of 45 ft3  of  sample volume  is required  for the
      determination  of  the  Destruction  and  Removal   Efficiency  (ORE)  of
      formaldehyde from incineration systems (45 ft3 is  equivalent  to  one hour
      of sampling at  0.75 dscf).  Additional  sample volume shall be collected as
      necessitated by the capacity  of the DNPH reagent and analytical  detection
      limit constraints.  To determine the minimum  sample volume required, refer
      to sample calculations in Sec. 10.0.

            6.3.5  Determine the total  length of sampling time needed to obtain
      the  identified minimum  volume  by comparing   the anticipated  average
      sampling rate with  the volume requirement.  Allocate the same time to all
      traverse points defined by  EPA Method 1.  To  avoid  timekeeping errors, the
      length of time  sampled at each traverse point should  be an  integer or an
      integer plus 0.5 min.

            6.3.6    In  some  circumstances   (e.g.,  batch  cycles)  it  may  be
      necessary to sample for shorter times at the  traverse  points and to obtain
      smaller gas-volume  samples.   In these cases,  careful documentation must be
      maintained in order to allow accurate  calculation  of  concentrations.

      6.4  Preparation of Collection Train

            6.4.1  During preparation  and assembly of the sampling  train, keep
      all openings where contamination can occur  covered with  Teflon®  film or
      aluminum foil until just prior to assembly or until sampling  is  about to
      begin.

NOTE: Appendix A at the end  of  this  procedure contains guidance on the  addition
      of a filter as  a check on the  survival of particulate material  through the
      impinger system.  This filter can  be added  to  the impinger train either
      after the second impinger or after the third impinger.

            6.4.2  Place 100 ml  of  cleaned DNPH solution in each of  the first
      two impingers,  and  leave  the third impinger empty.   If additional  capacity
      is required for high expected  concentrations of formaldehyde in the stack
      gas, 200 ml of  DNPH per  impinger  may be used or additional  impingers may
      be used for sampling.  Transfer approximately 200  to 300 g of pre-weighed
      silica gel  from  its  container to  the  fourth  impinger.  Care should be
      taken to ensure that the  silica gel is not entrained and carried  out from
      the impinger during sampling.  Place the silica gel container in  a clean
      place for later use in the  sample recovery.  Alternatively,  the weight of
                                   0011 - 9                         Revision 0
                                                                  January 1995

-------
      the silica gel plus impinger may be determined to the nearest  0.5  g  and
      recorded.

            6.4.3  With  a glass  or quartz liner, install the  selected  nozzle
      using a Viton-A 0-ring when stack temperatures are less than 260°C (500°F)
      and a woven glass-fiber gasket when temperatures are higher.   See  APTD-
      0576 (Rom, 1972)  for details.  Other connecting systems  utilizing  either
      316 stainless steel or Teflon®  ferrules may be used.  Mark the probe with
      heat-resistant tape or by some  other method to denote the proper distance
      into the stack or duct  for  each sampling  point.

            6.4.4  Assemble the train as shown in  Figure 1.  During assembly, do
      not  use any  silicone  grease  on ground-glass  joints   upstream of  the
      impingers.   Use  Teflon® tape,  if required.   A very  light coating  of
      silicone grease  may be used on ground-glass  joints  downstream of  the
      impingers, but the silicone grease should be limited to the outer portion
      (see APTD-0576)  of the  ground-glass  joints to minimize  silicone  grease
      contamination.   If necessary,  Teflon® tape may  be  used to  seal  leaks.
      Connect all temperature  sensors to an appropriate potentiometer/ display
      unit.  Check all  temperature  sensors at ambient temperature.

            6.4.5  Place crushed  ice all  around the  impingers.

            6.4.6   Turn on and  set  the probe heating  system at  the desired
      operating temperature.   Allow time for the  temperature  to stabilize.

      6.5  Leak-Check Procedures

            6.5.1  Pre-test Leak  Check

                  6.5.1.1 After  the  sampling train has been assembled, turn on
            and  set  the  probe   heating  system  at  the  desired  operating
            temperature.  Allow  time for  the temperature to stabilize.   If a
            Viton-A 0-ring or other leak-free connection  is  used in assembling
            the probe  nozzle  to  the  probe liner, leak-check the train  at  the
            sampling site by plugging the nozzle and  pulling a 381 mm Hg (15 in.
            Hg) vacuum.

NOTE: A lower vacuum may be used,  provided that  the lower vacuum is not exceeded
      during the test.

                  6.5.1.2  If an asbestos  string is used, do  not  connect  the
            probe to the train during the leak check.  Instead, leak-check the
            train by first attaching  a  carbon-filled leak check impinger to the
            inlet and  then plugging  the inlet  and pulling a 381  mm Hg (15  in.
            Hg) vacuum.  (A lower vacuum may be  used  if this lower vacuum is not
            exceeded during the test.)   Then  connect the probe to the train and
            leak-check  at about  25 mm Hg  (1  in. Hg) vacuum.   Alternatively,
            leak-check the probe  with the rest of the  sampling train in one step
            at 381 mm Hg (15 in.  Hg) vacuum.   Leakage rates  no greater than 4%


                                   0011  -  10                         Revision 0
                                                                  January 1995

-------
            of the average sampling  rate or less than or equal  to 0.00057 m3/roiri
            (0.02 cfm), whichever is less, are acceptable.

                  6.5.1.3    The  following  leak  check  instructions  for  the
            sampling train described in APTD-0576  and APTD-0581 may be helpful.
            Start the  pump  with  the fine-adjust valve  fully  open  and coarse-
            adjust valve  completely closed.   Partially open  the  coarse-adjust
            valve  and  slowly close  the   fine-adjust  valve  until   the desired
            vacuum  is  reached.   Do not  reverse direction of  the  fine-adjust
            valve, as liquid will back up into the train.  If the desired vacuum
            is exceeded, either  perform the leak check  at this higher vacuum or
            end the leak check,  as shown  below, and start over.

                  6.5.1.4  When the leak check is completed, first slowly remove
            the plug from the inlet  to the probe.  When  the vacuum drops to 127
            mm (5 in.)  Hg  or less,  immediately close  the coarse-adjust valve.
            Switch off the pumping system  and  reopen the fine-adjust valve.  Do
            not reopen the fine-adjust valve  until the coarse-adjust valve has
            been closed to prevent the liquid  in the  impingers from being forced
            backward into the sampling line and  silica gel from being entrained
            backward into the third  impinger.

            6.5.2  Sampling Run  Leak Check

                  6.5.2.1   If,  during the  sampling  run,  a  component  change
            (i.e., impinger) becomes necessary,  a  leak check shall  be conducted
            immediately after the interruption of sampling and  before the change
            is made.  The leak check shall be  done  according  to the procedure
            described in Sec. 6.5.1, except that  it shall be  done  at  a vacuum
            greater than or equal to the maximum value  recorded  up to that point
            in the test.   If the  leakage  rate is found to be  no greater than
            0.00057  m3/min  (0.02 cfm)  or 4%  of  the  average sampling  rate
            (whichever  is less),  the  results  are  acceptable.   If  a  higher
            leakage rate is obtained,  the  tester must  void the sampling run.

NOTE: Any correction of the sample volume  by  calculation reduces  the integrity
      of the pollutant concentration data  generated and must  be avoided.

                  6.5.2.2   Immediately  after  a  component  change   and  before
            sampling is  reinitiated, a leak  check similar to  a pre-test  leak
            check must also be conducted.

            6.5.3   Post-test Leak Check  - A leak  check  is  mandatory at  the
      conclusion of each  sampling run.  The leak  check shall  be done  with the
      same procedures as the pre-test leak check, except  that the post-test leak
      check shall  be conducted at a  vacuum greater than or equal to  the maximum
      value reached during the sampling run.   If the  leakage rate is found to be
      no greater than 0.00057 m3/min (0.02 cfm) or  4%  of  the  average  sampling
      rate (whichever  is less),  the results  are  acceptable.   If,   however,  a
                                  0011 - 11                         Revision  0
                                                                  January 1995

-------
higher leakage rate is obtained,  the tester shall record the leakage rate
and void the sampling run.

6.6  Sampling Train Operation

      6.6.1  During the sampling run,  maintain  an isokinetic sampling rate
to within 10% of true  isokinetic,  below  28 L/min  (1.0  cfm).   Maintain a
temperature around the probe of 120°C (248°  ±25°F).

      6.6.2  For each run, record the data on  a data  sheet such as the one
shown in Figure 2.   Be sure to record the initial  dry-gas meter reading.
Record  the  dry-gas  meter readings  at  the  beginning  and  end of  each
sampling time increment,  when changes in flow rates  are made, before and
after each leak check, and when sampling is  halted.   Take other readings
required by Figure 2 at least once at each sample point during each time
increment  and  additional  readings  when significant  adjustments  (20%
variation in velocity head readings) necessitate additional adjustments in
flow rate.  Level and zero the manometer.   Because the manometer level and
zero may drift  due to vibrations  and  temperature  changes,  make periodic
checks during the  traverse.

      6.6.3   Clean  the  stack access  ports   prior  to the  test run  to
eliminate the chance of sampling deposited material.  To begin sampling,
remove the nozzle  cap, verify  that the  filter and probe  heating systems
are at the specified temperature, and verify that the pitot tube and probe
are properly positioned.   Position  the nozzle  at the  first traverse point,
with the tip pointing directly  into the gas stream.   Immediately start the
pump and adjust the flow  to  isokinetic conditions.  Nomographs, which aid
in the rapid adjustment of the  isokinetic  sampling rate without excessive
computations, are  available.  These nomographs are designed for use when
the  Type S  pitot   tube  coefficient  is  0.84  ±0.02 and  the  stack  gas
equivalent density (dry molecular  weight) is  equal  to  29 ±4.  APTD-0576
details  the  procedure  for  using  the  nomographs.    If  the  stack  gas
molecular weight  and the pitot  tube  coefficient  are outside  the  above
ranges, do not  use the nomographs  unless  appropriate  steps  are taken to
compensate for the deviations.

      6.6.4    When  the  stack is  under  significant  negative  pressure
(equivalent to  the height of the  impinger stem),  take  care  to close the
coarse-adjust valve before inserting  the  probe into  the stack in order to
prevent liquid from backing up through the train.  If necessary, the pump
may be turned on with the coarse-adjust valve closed.

      6.6.5  When the probe is in  position, block off the openings around
the probe and stack  access port to prevent nonrepresentative dilution of
the gas stream.

      6.6.6  Traverse the stack cross section, as required by EPA Method
1, being  careful not to  bump the  probe  nozzle into  the stack walls when
sampling  near  the  walls  or  when  removing or  inserting  the probe through
                             0011  -  12                        Revision 0
                                                            January 1995

-------
      the access port, in order to minimize the chance of extracting deposited
      material.

            6.6.7  During the test run, make  periodic  adjustments  to keep the
      temperature around the probe at the proper levels.   Add more  ice and,  if
      necessary, salt, to maintain a temperature of less than  20°C (68°F) at the
      silica gel outlet.  Also,  periodically  check the level and  zero  of the
      manometer.

            6.6.8  A  single  train  shall  be used for the  entire  sampling  run,
      except in cases  where  simultaneous  sampling is required in  two  or  more
      separate ducts  or at two or more different locations  within the same duct,
      or in cases where equipment failure necessitates a  change  of  trains.  An
      additional train or additional  trains may also be used  for  sampling  when
      the capacity of a single train  is exceeded.

            6.6.9   When two  or  more  trains  are  used,  separate  analyses  of
      components from each train shall be  performed.   If  multiple  trains  have
      been used because the  capacity  of a single train would be exceeded, first
      impingers from each train may be  combined, and second impingers from each
      train may be combined.

            6.6.10  At the  end of the sampling run, turn  off  the  coarse-adjust
      valve, remove the  probe and nozzle  from  the stack, turn off  the pump,
      record the  final  dry gas  meter reading, and  conduct  a post-test  leak
      check.  Also, leak check the  pitot lines as described in  EPA  Method  2.
      The lines  must pass this leak check in order to validate the velocity-head
      data.

            6.6.11  Calculate percent isokineticity (see Method 2) to determine
      whether the run was valid or another test should be made.
7.0  SAMPLE RECOVERY AND PREPARATION FOR ANALYSIS

      7.1  Preparation

            7.1.1   Proper cleanup procedure  begins  as soon  as  the  probe  is
      removed from the stack at the end of the  sampling  period.  Allow the probe
      to cool.   When the probe can  be  handled  safely, wipe  off  all  external
      particulate matter near the tip of the probe nozzle  and place a cap over
      the tip to prevent losing or gaining particulate  matter.  Do not cap the
      probe tip tightly  while  the sampling train is cooling  because  a vacuum
      will  be  created,  drawing  liquid  from  the impingers  back  through  the
      sampling train.

            7.1.2  Before moving the  sampling train to the cleanup site, remove
      the probe from the sampling train and cap  the  open outlet,  being careful
      not to lose any condensate  that might be  present.   Remove  the umbilical
      cord from the last impinger and cap the  impinger.  If a flexible line  is
      used, let any  condensed water or liquid drain into the impingers.  Cap off

                                  0011  -  13                         Revision 0
                                                                  January 1995

-------
any open  impinger inlets and  outlets.   Ground glass  stoppers,  Teflon®
caps, or caps of other inert materials may be used to seal all openings.

      7.1.3  Transfer the probe and impinger assembly to an area that is
clean and  protected  from wind  so  that  the chances  of  contaminating or
losing the sample are minimized.

      7.1.4  Inspect  the  train before and during  disassembly, and note any
abnormal conditions.

      7.1.5  Save a portion of all  washing solution (methylene chloride,
water) used  for  cleanup  as a  blank.   Transfer  200 ml  of each solution
directly from the wash  bottle  being used and place  each  in  a separate,
pre-labeled sample container.

7.2  Sample Containers

      7.2.1  Container 1  -  Probe and  Impinger Catches.  Using a graduated
cylinder, measure to  the  nearest ml, and record the volume  of the solution
in the first three impingers.   Alternatively, the  solution may be weighed
to  the  nearest  0.5  g.    Include  any  condensate   in  the  probe  in  this
determination. Transfer  the impinger solution from the graduated cylinder
into the amber flint glass bottle.   Taking care that dust on the outside
of  the  probe  or  other exterior  surfaces  does  not get  into  the sample,
clean all  surfaces to which the sample is exposed (including the probe
nozzle,   probe  fitting,   probe  liner,   first   impinger,   and  impinger
connector) with methylene chloride.  Use less than 500 ml for the entire
wash  (250  ml would  be better,  if possible).   Add the washings  to the
sample container.

            7.2.1.1   Carefully  remove  the probe  nozzle  and  rinse the
      inside surface with methylene  chloride from a  wash  bottle.   Brush
      with a Teflon® bristle  brush,  and rinse  until  the  rinse shows no
      visible particles or yellow color, after which make a final rinse of
      the  inside surface.   Brush  and  rinse  the inside  parts  of the
      Swagelok fitting with methylene chloride  in  a similar way.

            7.2.1.2  Rinse the probe 1iner with methylene chloride.  While
      squirting the  methylene  chloride  into  the upper end of the probe,
      tilt and rotate the probe  so that all  inside  surfaces will be wetted
      with methylene chloride.   Let the  methylene  chloride drain from the
      lower end  into  the sample container.   The tester may use a funnel
      (glass or polyethylene)  to aid in  transferring the liquid washes to
      the container.   Following the rinse with a Teflon® brush.  Hold the
      probe in an  inclined  position,  and  squirt methylene chloride into
      the  upper  end  as  the probe brush is being  pushed with  a twisting
      action through the probe.   Hold the sample container underneath the
      lower end of the probe, and catch any  methylene  chloride, water, and
      particulate matter that  is brushed from the probe.   Run the brush
      through the  probe  three  times  or more.   With  stainless  steel  or
      other metal probes, run  the  brush through in the above prescribed

                             0011 - 14                        Revision 0
                                                            January 1995

-------
      manner at least six times since there may be small crevices in which
      particulate matter can be entrapped.  Rinse the  brush with methylene
      chloride or water, and quantitatively collect these washings in the
      sample container.   After the  brushings,  make  a final  rinse  of the
      probe as described  above.

NOTE: Between  sampling  runs, brushes must  be  kept  clean and  free  from
      contamination.

            7.2.1.3  Rinse the inside surface of each of the first three
      impingers (and connecting tubing) three separate times.  Use a small
      portion of methylene chloride  for each rinse, and brush each surface
      to which sample is  exposed with  a  Teflon® bristle  brush  to  ensure
      recovery of fine particulate matter.  Water will  be required for the
      recovery of the impingers in  addition to the specified quantity of
      methylene  chloride.   There  will   be  at  least  two  phases  in  the
      impingers.    This  two-phase  mixture  does  not pour  well,  and  a
      significant amount of the impinger  catch  will be left on the walls.
      The use of water as a  rinse makes the recovery quantitative.  Make
      a final  rinse of each  surface  and of the brush,  using both methylene
      chloride and water.

            7.2.1.4  After all methylene chloride and water washings and
      particulate  matter have been  collected  in the  sample  container,
      tighten the  lid so  that  solvent, water,  and DNPH  reagent will  not
      leak out when the  container is snipped to the laboratory.  Mark the
      height of the fluid level to determine whether leakage occurs during
      transport.    Seal  the  container  with Teflon® tape.    Label  the
      container clearly to identify its contents.

            7.2.1.5   If  the first two  impingers  are  to  be  analyzed
      separately  to check for breakthrough,  separate the  contents  and
      rinses of the two impingers into individual containers.   Care must
      be taken to avoid  physical  carryover from the first impinger to the
      second.    The formaldehyde  hydrazone  is  a solid which  floats  and
      froths on top of the impinger solution.  Any physical  carryover of
      collected  moisture  into  the  second  impinger  will  invalidate  a
      breakthrough assessment.

      7.2.2  Container 2 - Sample Blank.   Prepare  a sample blank by using
an amber flint glass  container  and adding  a volume  of DNPH  reagent  and
methylene chloride equal to the total volume in Container 1.  Process the
blank in the same manner  as Container 1.

      7.2.3  Container 3  - Silica Gel.   Note the color of the indicating
silica gel  to determine  whether  it  has been completely spent  and  make a
notation of its condition.  The impinger  containing the silica gel  may be
used as a sample  transport  container with both  ends  sealed  with tightly
fitting caps or plugs.  Ground-glass stoppers or Teflon® caps may be used.
The  silica  gel  impinger  should  then  be  labeled, covered with aluminum
foil, and packaged on ice for transport to the laboratory.  If the  silica

                             0011  -  15                         Revision 0
                                                            January 1995

-------
      gel is removed from the impinger,  the tester may use a funnel to pour the
      silica gel  and a  rubber policeman to  remove the  silica  gel  from  the
      impinger.   It  is  not  necessary  to  remove  the  small  amount of  dust
      particles  that  may adhere  to the  impinger  wall  and  are  difficult  to
      remove.  Since the  gain in weight is to  be used for moisture calculations,
      do not  use water  or  other liquids to  transfer  the silica gel.    If  a
      balance is available  in  the field, the spent silica gel  (or silica gel
      plus impinger) may be weighed to the nearest 0.5  g.

            7.2.4   Sample containers  should  be  placed  in a  cooler,  cooled  by
      although  not  in contact  with ice.   Sample  containers  must  be  placed
      vertically and,  since they  are  glass, protected  from  breakage  during
      shipment.   Samples should  be cooled  during  shipment so  they will  be
      received cold at the laboratory.


8.0  CALIBRATION

      8.1  Probe Nozzle - Probe nozzles shall  be calibrated before their initial
use in the field.  Using  a micrometer,  measure the  inside diameter  of the nozzle
to the nearest  0.025 mm (0.001 in.). Make measurements at three separate places
across the diameter and  obtain  the average of  the measurements.  The difference
between the high and low numbers shall  not exceed 0.1 mm (0.004 in.).  When the
nozzles become nicked or corroded,  they shall  be replaced and calibrated before
use.  Each nozzle must be permanently  and uniquely identified.

      8.2   Pitot Tube -  The Type S pitot  tube  assembly shall  be calibrated
according to the procedure  outlined in Sec.  4 of EPA Method 2,  or assigned  a
nominal coefficient of 0.84  if  it  is not visibly  nicked or  corroded and if it
meets design and intercomponent spacing specifications.

      8.3  Metering System

            8.3.1   Before  its  initial  use  in the field,  the  metering  system
      shall  be  calibrated  according to  the  procedure  outlined  in APTD-0576.
      Instead  of physically  adjusting  the   dry-gas  meter  dial   readings  to
      correspond to the wet-test meter readings,  calibration factors may be used
      to  correct the gas  meter  dial  readings  mathematically to  the  proper
      values.   Before calibrating  the  metering system,  it  is suggested  that  a
      leak check be conducted.   For metering  system having diaphragm pumps, the
      normal leak check  procedure will  not detect leakages  within  the pump.  For
      these cases,  the following leak  check procedure will apply:  make a ten-
      minute calibration run at  0.00057  m3/min (0.02 cfm).   At the end  of the
      run,  take the  difference of the measured wet-test  and  dry-gas  meter
      volumes and divide the difference  by 10 to  get the  leak  rate.  The leak
      rate should not exceed 0.00057 m3/min  (0.02  cfm).

            8.3.2  After each field use, check the calibration of the metering
      system  by performing  three  calibration runs  at  a single  intermediate
      orifice setting (based on the  previous field test).   Set the  vacuum at the


                                   0011  - 16                         Revision 0
                                                                  January 1995

-------
      maximum  value  reached during  the  test series.   To adjust  the  vacuum,
      insert a valve between the wet-test meter  and  the inlet of the metering
      system.  Calculate the average  value  of the  calibration factor.   If the
      calibration has changed  by more than  5%,  recalibrate  the meter over the
      full range of orifice settings, as outlined in APTD-0576.

            8.3.3  Leak Check of Metering System - The portion of the sampling
      train from the pump  to the orifice meter  (see  Figure  1) should be leak-
      checked prior to initial  use and after each shipment.   Leakage after the
      pump will result in less volume being recorded than is actually sampled.
      Use the  following  procedure:   Close  the  main  valve  on the  meter  box.
      Insert a  one-hole  rubber stopper with  rubber tubing attached  into the
      orifice exhaust pipe.   Disconnect  and vent the low side of the  orifice
      manometer.  Close  off the low side orifice tap.  Pressurize the system to
      13  -  18  cm  (5  -  7  in.) water column  by blowing  into  the rubber  tubing.
      Pinch off  the  tubing and observe  the manometer  for  1  min.  A  loss  of
      pressure on the manometer indicates a leak in the meter box.  Leaks must
      be corrected.

NOTE: If the dry-gas-meter coefficient values obtained before and after a test
      series differ by greater  than 5%, either the test series  must be voided or
      calculations  for  test series  must  be  performed  using  whichever meter
      coefficient value (i.e.,  before or after)  gives the lower value of total
      sample volume.

      8.4  Probe Heater - The probe heating  system must be calibrated before its
initial  use  in the  field  according  to the procedure  outlined  in  APTD-0576.
Probes  constructed  according  to  APTD-0581 need  not  be  calibrated   if  the
calibration curves in APTD-0576 are used.

      8.5   Temperature  Gauges -   Each  thermocouple  must  be  permanently  and
uniquely marked on the casting.  All mercury-in-glass reference  thermometers must
conform  to  ASTM E-l  63C or  63F  (American  Society for  Testing and  Materials)
specifications.  Thermocouples should be calibrated in the laboratory with and
without the use of extension leads.   If extension leads are used in the field,
the thermocouple readings at the ambient air temperatures, with and without the
extension lead, must be  noted and recorded.  Correction  is necessary if the use
of an extension lead produces a change greater than 1.5%.

            8.5.1    Impinger  and  Dry-gas  Meter  Thermocouples  -  For  the
      thermocouples  used  to measure  the  temperature of  the  gas leaving  the
      impinger  train, a  three-point  calibration at  ice water,  room air,  and
      boiling water temperatures is necessary. Accept the thermocouples only if
      the readings at all three temperatures agree to ±2°C  (3.6°F) with those of
      the absolute value of the reference thermometer.

            8.5.2  Probe  and Stack  Thermocouple - For the thermocouples used to
      indicate the probe and stack  temperatures, a  three-point calibration  at
      ice water, boiling water, and hot oil  bath temperatures must be performed.
      Use of a point at room air temperature  is  recommended.   The thermometer
      and thermocouple must  agree  to within  1.5%  at each of the calibration

                                   0011 - 17                        Revision 0
                                                                  January 1995

-------
      points.  A  calibration  curve  (equation)  may be constructed  (calculated)
      and the data extrapolated to cover the  entire temperature range  suggested
      by the manufacturer.

      8.6  Barometer - Adjust the barometer initially and before each test series
to agree  to within ±2.5  mm  Hg (0.1 in.  Hg)  of the mercury  barometer or  the
corrected barometric pressure value reported by  a nearby National Weather Service
Station (same altitude above sea level).

      8.7  Triple-beam or  Electronic Balance - Calibrate the balance before each
test series, using Class  S standard weights.  The weights must be  within ±0.5%
of the standards, or the  balance must be  adjusted to meet these limits.


9.0  CALCULATIONS

      Perform calculations, retaining at  least one extra decimal figure  beyond
that of the acquired data.  Round off figures after final calculations.

      9.1  Total  Formaldehyde - Determine the total formaldehyde in mg,  using  the
following equation:

                                           [g/mole aldehyde]
Total mg formaldehyde = Cd x V x DF  x  	 x  103  mg/^ug
                                        [g/mole DNPH derivative]

      where:

         Cd = measured concentration of DNPH-formaldehyde derivative,  ng/ml
          V = organic extract volume, mL
         DF = dilution factor

      9.2  Formaldehyde Concentration In Stack  Gas  - Determine the  formaldehyde
concentration in  the stack gas using the  following equation:

            Cf  =  K [total  formaldehyde,  mg] / Vm(std)

      where:

            K  =35.31  ft3/")3 if Vm(std) is expressed in  English  units
               =  1.00 m3/m3 if Vm(std,  is  expressed in metric units
        Vm(stdi  =  volume of gas sample as measured by dry gas meter,  corrected to
                 standard  conditions, dscm (dscf)

      9.3  Average Dry Gas Meter Temperature and Average Orifice Pressure Drop
are obtained from the data sheet.

      9.4  Dry  Gas Volume  - Calculate Vm(std) and adjust for leakage,  if necessary,
using the equation in Sec. 6.3 of EPA Method 5.
                                   0011 - 18                         Revision  0
                                                                   January  1995

-------
      9.5  Volume of Water Vapor and Moisture Content  -  Calculate  the  volume  of
water vapor and moisture content from Equations 5-2 and 5-3 of  EPA Method  5.


10.0  DETERMINATION OF VOLUME TO BE SAMPLED

      To determine the minimum sample volume to be  collected,  use  the  following
sequence of equations.

      10.1    From  prior  analysis   of  the  waste   feed,  the  concentration  of
formaldehyde (FORM) introduced into the combustion  system can be calculated.  The
degree  of destruction  and  removal  efficiency that  is  required is  used  to
determine the maximum amount  of FORM allowed to be present in the effluent.  This
amount may be expressed as:

            Max FORMi  Mass  =  [(WF)(FORM, cone) (100  - %DRE)]  / 100

      where:

         WF = mass flow rate  of waste feed per h, g/h (Ib/h)
      FORM,  = concentration  of FORM (wt %) introduced into the  combustion
              process
        ORE = percent Destruction and Removal  Efficiency required
   Max FORM = mass flow rate  (g/h [lb/h]) of  FORM emitted from  the combustion
              sources

      10.2  The average discharge concentration of the FORM  in the effluent gas
is determined  by  comparing  the  Max FORM with the  volumetric  flow rate being
exhausted from the source.  Volumetric flow rate data  are available as a  result
of preliminary EPA Method  1-4 determinations:

            Max FORMi  cone  =  [Max FORM, Mass]  / DVeff(std)

      where:

         DVeff(stdi    = volumetric flow rate  of  exhaust gas, dscm  (dscf)
       FORM; cone   =  anticipated  concentration of  the  FORM in the exhaust  gas
                     stream,  g/dscm  (Ib/dscf)

      10.3  In making this calculation, it is recommended that  a  safety  margin
of at least ten be included.
                    [LDLFORM x 10] / [FORMi cone] = Vtbc

      where:

            LDLFORM  =  detectable  amount  of FORM in  entire sampling train
            Vtbc     =  minimum  dry standard volume  to be collected at dry-gas
                      meter
                                   0011  -  19                         Revision 0
                                                                  January 1995

-------
      10.4  The following analytical  detection limits and DNPH Reagent Capacity
(based on a total  volume of 200  mL  in two impingers) must also be considered in
determining a volume to be sampled.


11.0  QUALITY CONTROL

      11.1  Sampling  - See EPA Manual 600/4-77-027b for Method 5 quality control.

      11.2  Analysis -  The  quality assurance  program  required  for this method
includes the analysis of field and method blanks,  procedure validations, analysis
of field spikes, and analysis of reagent checks.  The assessment of combustion
data and positive  identification and  quantitation of formaldehyde are dependent
on the integrity of the samples received and the precision and accuracy of the
analytical  methodology.    Quality  Assurance  procedures for  this method  are
designed to monitor the  performance of the analytical methodology and to provide
the required information to take corrective action if problems are observed in
laboratory operations or in field sampling activities.

            11.2.1   Field Blanks  -  Field blanks  must  be  submitted  with  the
      samples collected  at  each sampling site.   The  field  blanks  include  the
      sample bottles  containing  aliquots of sample  recovery solvents, methylene
      chloride and water, and unused DNPH reagent.  At a minimum,  one complete
      sampling train will be assembled in the field staging area,  taken to the
      sampling area,  and  leak-checked  at the  beginning  and  end  of the testing
      (or for the  same total number of times as the actual  sampling train).  The
      probe of the blank train must be heated during the  sample test.  The train
      will be recovered as if it were an actual  test sample.  No gaseous sample
      will be passed through the Blank sampling train.

            11.2.2  Method Blanks - A method blank must be  prepared for each set
      of analytical operations,  to evaluate contamination and artifacts that can
      be  derived  from  glassware,  reagents,  and  sample  handling  in  the
      laboratory.

            11.2.3   Field  Spikes  - A field spike  is  performed  by introducing
      200 nl of the  Field Spike Standard into an impinger containing 200 mL of
      DNPH solution.  Standard impinger recovery procedures are followed and the
      field spike sample is returned to the laboratory for analysis.  The field
      spike is used  as  a  check  on  field handling  and  recovery procedures.   An
      aliquot  of  the field spike  standard  is retained  in  the  laboratory  for
      derivatization and comparative analysis.

            11.2.4   DNPH Reagent  Checks  - An  aliquot  of  the  extracted  DNPH
      reagent is prepared and analyzed  according to the  procedure in Sec. 5.4.3
      to ensure that the background in  the  reagent  is  acceptable for field use.
                                   0011  -  20                        Revision 0
                                                                  January 1995

-------
12.0  METHOD PERFORMANCE

      12.1   Method  performance evaluation  - The  following  expected  method
performance parameters for precision, accuracy,  and detection limits are provided
in Table 3.
13.0  REFERENCES

1.    U.S.  Environmental  Protection Agency, 40  CFR Part 60  Appendix  A, Test
      Methods.

2.    Martin,  R.M.,   "Construction   Details   of  Isokinetic  Source-Sampling
      Equipment", U.S. Environmental  Protection Agency, Research Triangle Park,
      NC, Air Pollution Technical Document (APTD) 0581, April 1971.

3.    Rom, J.J.,  "Maintenance, Calibration, and Operation of  Isokinetic Source
      Sampling  Equipment",  U.S.   Environmental  Protection   Agency,  Research
      Triangle Park,  NC,  Air Pollution Technical Document  (APTD)  0576,  March
      1972.

4.    Annual  Book  of ASTM Standards.  Part  26. Gaseous Fuels; Coal  and  Coke;
      Atmospheric Analysis.  American Society for Testing and  Materials (ASTM),
      Philadelphia, PA, 1974, pp. 617-622.
                                  0011  - 21                         Revision 0
                                                                  January 1995

-------
                                   TABLE  1
                 APPROXIMATE AMOUNT OF CRYSTALLINE DNPH USED
                        TO  PREPARE A SATURATED SOLUTION
     Amount of Moisture in DNPH           Weight Required per 8 L of Solution
          10 weight  percent                                31 g
          15 weight  percent                                33 g
          30 weight  percent                                40 g
                                    TABLE  2
               INSTRUMENT DETECTION LIMITS AND REAGENT CAPACITY
                          FOR FORMALDEHYDE ANALYSIS1
Analyte                      Detection Limit,  ppbv2    Reagent Capacity, ppmv
Formaldehyde
Acetaldehyde
Acrolein
Acetone/Propi onal dehyde
Butyraldehyde
Methyl ethyl ketone
Valeral dehyde
Isovaleraldehyde
Hexal dehyde
Benzal dehyde
o-/m-/p-Tolualdehyde
Dimethyl benzal dehyde
1.8
1.7
1.5
1.5
1.5
1.5
1.5
1.4
1.3
1.4
1.3
1.2
66
70
75
75
79
79
84
84
88
84
89
93
1 Oxygenated compounds in addition to formaldehyde are included for comparison
with formaldehyde; extension of the methodology to other compounds is possible.

2 Detection limits are determined in solvent.   These values therefore represent
the optimum capability of the methodology.
                                   0011 - 22                        Revision 0
                                                                  January 1995

-------
                                    TABLE  3
                 EXPECTED METHOD PERFORMANCE FOR FORMALDEHYDE
                                                                           3
Parameter          Precision1             Accuracy2         Detection Limit


Matrix:  Dual      ±15% RPD                 ±20%            1.4 x 10'10 lb/ft3
  trains                                                        (1.8 ppbv)


1  Relative percent difference limit for dual  trains.
2  Limit for field spike recoveries.
3  The lower reporting limit having less than  1% probability of false positive
  detection.
                                  0011  - 23                         Revision 0
                                                                  January 1995

-------
O ^
                                                                                                                                    i
 

 o
1 >•
 LU
 O
 (M

  i
 O
 O

-------
i
u
^*
c
e
£
E





i
c
•:-
*
i





      2    £
       t.
       Ol
               VI    Ol
               v>    fc.
               O)    3
               k.   *J
                                         Ol
                                 1.0)
           tD.   V»   ^   4->   -^4)   C
                 f   **    c   f— E   ••-
            u   O    O>   0)   «• «f   <->
      H-   ^-   Z    C   T3   O-^   «   0>
            fc.         O)   i—>     O   Ol   **

       c    o>   o)         oi   o>0)        oe
       O>    E   E    Ol   >—   Mr-   W
      •»-    o   5   A    M   UN   £   j<
      JO    i-   V»    O    M   Ol N   O   !•    O   ««   F—
       g    ««   VI    t.    O   >O   t-   Ol    fc.  *»   •*•
      <   CD   <   a.   z   «z   a.   —i   o.  c/>   u.
 M
•^

 Ol
*>    0)
 M    U
JC    3

 U    V)
 Ol    Ol
 C    fc.   O
•>-   a.   z

      u   t.
 01   •»-   0)
.a   *»   **
Q
CJ _l
>—i Q-



"I

    O
                                             c
                                             o
                                              u
                                              o
                                             to
                                              o

                                              u
                                       X
                                       O   •>
                                      CO   Z
                                                         J3 Ol
                                                     fc.    3-^
                                                     O   H* W
                       O>
            u   o>   4->    e
            O   fi.   ra    3
           —i   O   o   a:
                                    t—    1.   I-    O
                                     Q.   O>   Ol    «
                                     E   4J   *J   U.
                                     ™    a)   01
                                    uo   ac   z   o
   o**-

   •»- o
>*- fc-
c cnu.
t>4-> ••---»
V> 19  Q.
«o 01 <» E o
u fc. iT"
0> 0) ••
^-"o oio
u-zi— •
>» OILL.
u »— •
OIO *>— '
t- I. 30
O-t-» Ol O*
M Ol «-••
«/> •£ 4->U-
tA) •
«/» I—-'
<« CO
cat— is •"•«•
(0 « On
15
oi c —  U.
4-> O)
^^i
U 0 3
t— cxz








































































QJ
>
D>










ro
O
















!"
Average
                                                                                                                                 c en
                                                                                                                                 o ^
                                                                                                                                 O)  3
                                                                                                                                 Q£  C
                                                                                                                                 CNJ


                                                                                                                                  I

-------
                                  APPENDIX A

           ADDITION  OF A  FILTER TO THE  FORMALDEHYDE SAMPLING TRAIN


      As a check on  the survival  of  participate  material  through the impinger
system, a  filter  can be added to the  impinger train either after  the  second
impinger or after the third impinger.  Since the  impingers are  in an ice bath,
there is no reason to heat the filter at this point.

      Any suitable medium  (e.g., paper, organic  membrane)  may  be used  for the
filter if the material conforms to the following  specifications:

   1) The filter has  at  least  95% collection efficiency (<5% penetration) for 3
      /xm dioctyl phthalate  smoke  particles.   The filter  efficiency  test shall
      be conducted in accordance with ASTM standard method D2986-71.   Test data
      from the  supplier's  quality   control  program  are  sufficient for  this
      purpose.

   2) The filter has a low aldehyde  blank value  (<0.015 mg formaldehyde/cm2 of
      filter area).   Before the test  series, determine the average formaldehyde
      blank  value  of at  least three  filters (from the  lot  to be used  for
      sampling)  using the applicable analytical procedures.

      Recover the exposed  filter into a  separate  clean container and return the
container over  ice  to the  laboratory  for analysis.    If  the  filter is being
analyzed for formaldehyde, the filter may  be recovered into a container or DNPH
reagent for shipment  back to the laboratory. If the filter  is being examined for
the presence of particulate material, the filter may  be recovered into  a clean
dry container and returned to the  laboratory.
                                   0011  -  26                         Revision 0
                                                                  January 1995

-------
                                 METHOD 0023A

            SAMPLING METHOD FOR POLYCHLORINATED DIBENZO-p-DIOXINS
                  AND POLYCHLORINATED DIBENZOFURAN EMISSIONS
                            FROM STATIONARY  SOURCES
1.0   SCOPE AND APPLICATION

      1.1  This  method  describes  the  sampling  procedure  to  be  used  for
determining stack  emissions of polychlorinated dibenzo-£-dioxins  (PCDDs)  and
polychlorinated dibenzofurans (PCDFs) from stationary sources.   The air sample
is collected and analyzed by the determinative portion of Methods 8280 or 8290
(Sees.  7.6  -  7.8).   This  method  describes the  procedures for  sampling  and
calculating results. This  method may  be modified to allow simultaneous sampling
and  analysis  for  polychlorinated  biphenyls  (PCBs),  polynuclear  aromatic
hydrocarbons  (PAHs),  or  semivolatile organic  compounds  (SVOCs).    However,
specific approval  is required for this modification,  and detailed modification
of the methodology is required.

          1.1.1    This  method  is a revision  of  EPA Air Method  23  (Ref.  12).

          1.1.2    The  surrogates and recovery  standards include the standards
      listed in Methods  8280 and 8290 (Table  1).   However,  analysts  should add
      the compounds  only one time, during sampling.

          1.1.3    The method refers to specific  techniques described in EPA Air
      Methods 1, 2  and 5.  Analysts should obtain copies of those methods prior
      to sampling.

      1.2  This  method   is  restricted to use by or  under  the  supervision  of
analysts experienced in the use of  air  sampling  methods and  the  analysis  of
PCDDs, PCDFs,  PCBs, PAHs,  and SVOCs from the components of Method 0010 trains.
Each analyst must demonstrate the ability to generate  acceptable  results with
this method.

      1.3  Safety - The  laboratory should develop a strict safety program for the
handling of PCDDs  and/or PCDFs.

          1.3.1    2,3,7,8-TCDD has been found to be acnegenic, carcinogenic, and
      teratogenic in laboratory animal  studies.  Other  PCDDs and  PCDFs containing
      chlorine atoms in  positions 2,3,7,8 are known  to  have toxicities comparable
      to that of 2,3,7,8-TCDD.   The analyst must be aware of the potential  for
      inhalation and ingestion.   It is recommended that such samples be processed
      in a  confined environment, such as a  hood  or a glove  box.   Personnel
      handling these types of samples  should wear masks fitted with  charcoal
      filters to prevent the  inhalation of airborne particulates.

          1.3.2    The toxicity  or carcinogenicity of each reagent used in this
     method is not precisely defined.  However, each  chemical should be treated
      as a potential  health  hazard,  and exposure to  these chemicals kept  to  a
     minimum. The laboratory is responsible for maintaining  a current awareness
      file of  OSHA  regulations  regarding  the safe handling of the chemicals
      specified in this  method.   A reference file of material safety data sheets

                                  0023A - 1                         Revision  1
                                                                  January 1995

-------
      should  be made available to  all  personnel  involved in  the  sampling and
      chemical  analysis of samples suspected to contain PCDDs/PCDFs.  Additional
      information on  laboratory safety  is given in References  4, 9,  10 and  11.


2.0   SUMMARY OF METHOD

      2.1   Gaseous and particulate PCDDs/PCDFs are isokinetically withdrawn from
an emission  source  and  collected in  a multicomponent  sampling  train.   The
collection components  consist of  the  front half glassware  surfaces (nozzle,
probe, and front  half filter holder),  the glass  fiber filter, the back half
glassware surfaces (back half filter holder and condenser  coil)  and the solid
sorbent (XAD-2®) module.

      2.2   Following  sampling  the glass collection  components are  rinsed.  The
PCDD/PCDF are then extracted from the front half rinses and filter and another
separate extraction  is performed on the XAD-2® and  back half rinses.

      2.3   The  filter  and  XAD-2®  extracts  are  then  analyzed  separately.
Surrogate  recoveries are  determined  for  both fractions.    The  analysis  is
performed using high resolution  gas  chromatography  (HRGC)  and high  resolution
mass spectrometry (HRMS), using the procedures  of Method 8290.


3.0   INTERFERENCES

      3.1   Polychlorinated biphenyls  (PCBs)  and polychlorinated diphenyl  ethers
(PCDPEs) may  interfere with low resolution gas chromatography/low resolution mass
spectroscopy PCDD/PCDF analyses,  since  these compounds produce many  of the same
nominal  masses  as   the  PCDDs  and  PCDFs.   However,  high  resolution  mass
spectrometric techniques, in combination with capillary gas chromatography, are
typically used to resolve PCDD/PCDFs from these analytical  interferences.

      3.2   Very high amounts  of other  organic  compounds  in the matrix will
interfere with the analysis.  Extensive column-chromatographic cleanup has been
introduced into  typical  HRGC/HRMS  analytical  methodology to  minimize  matrix
effects due to high  concentrations of organic compounds.

      3.3   Method  interferences  may  be  caused  by  contaminants   in solvents,
reagents,  glassware, and  other  sample processing  hardware.   All of these
materials must be routinely  demonstrated to be free  from  interferences under the
conditions of the analysis  by preparing and analyzing laboratory method blanks.

           3.3.1    Glassware must  be  cleaned  thoroughly before  using.   The
      glassware should be washed with laboratory detergent in  hot water followed
      by  rinsing with  tap  water  and  distilled  water.   The  glassware  may be
      cleaned  by baking  in  a glassware oven at 400°C  for at  least one hour.
      After the glassware has cooled, the glassware  should be  rinsed  three times
      with  methanol  and  three   times  with  methylene  chloride.    Volumetric
      glassware should  not  be heated  to  400°C.    Rather,  after washing and
      rinsing,  volumetric glassware  may be  rinsed  with methanol followed by
      methylene chloride  and allowed  to dry in air.
                                   0023A -  2                         Revision 1
                                                                  January 1995

-------
          3.3.2    The use of high purity reagents and solvents helps to minimize
      interference  problems  in sample analysis.


4.0   APPARATUS AND MATERIALS

      The  following  section  describes  all  the  sampling  equipment  and  the
associated performance specifications necessary to collect  a  gas sample from a
stationary source according to Method 0023.

      4.1  Sampling train  -  A  schematic diagram of the  sampling train  is shown
in Figure  1.  This train  configuration has been adapted from EPA Method 5 (40 CFR
Part 60  Appendix  A)  with  the   addition  of  condenser,   XAD-2®  trap   and
filtration-coil   connecting  glassware.   Sealing  greases must  not be  used  in
assembling the train.  Complete sampling systems are commercially available that
have been developed to meet all  the  EPA equipment design specifications.   The
following equipment is required.

          4.1.1    Nozzle  -  The nozzle should be made of quartz or borosilicate
      glass.  Stainless steel nozzles  should not be used.  The taper angle should
      be < 30°, with taper on the  outside to preserve a constant inside  diameter
      (ID).  The  nozzle ID  should be determined in order to sample isokinetically
      at a rate that allows collection of an adequate sample volume.  The minimum
      sample  volume should be  determined to allow  appropriate  detection limits
      to be achieved (see  Sec. 9.0).

          4.1.2    Probe liner - The sampling probe liner should be constructed
      of borosilicate or quartz glass  tubing.  The  typical outside diameter (OD)
      used  by sampling equipment  manufacturers is about 16 mm,  encased  in  a
      stainless  steel  sheath with  an  OD  of  25.4  mm.   Either  borosilicate  or
      quartz  glass  liners may be used for stack temperatures  up to about 480°C,
      but quartz glass liners should be  used at  higher stack temperature [480 to
      900°C].

          4.1.3    Probe  sheath  and  heating  system -  A stainless  steel  or
      equivalent probe sheath should be used to house the probe liner and heating
      system.  The  probe heating  system should  be  capable of maintaining probe
      gas temperatures at the probe exit of 120°C ± 14°C during sampling.  This
      temperature should  be verified by placing a thermocouple temperature sensor
      against  the outer  surface of the  probe  liner at least 2  feet upstream of
      the filter oven.  Temperature readings should be recorded during sampling.

          4.1.4    Glass cyclone - A glass  cyclone  may be used between the probe
      and filter holder for high particulate concentrations.  A cyclone,  if used,
      should  be rinsed and recovered with the front half of the train.

          4.1.5    Filter  holder  -  A filter holder of borosilicate glass with  a
      Teflon®  frit  filter support  should  be  used.   The  holder design  should
      provide  a  positive seal  against leakage  from  the outside or around  the
      filter.  The  holder  should  be durable,  easy  to  load, leak-free in normal
      applications,  and   is  positioned  immediately  following  the  probe  (or
      cyclone, if used) with the filter placed toward the flow.
                                  0023A - 3                         Revision  1
                                                                  January  1995

-------
     4.1.6   Filter heating system - Any heating system  may  be  used which
is  capable of  maintaining  the  filter holder  at  120°C ± 14°C  during
sampling.   Other  temperatures may  be  specified  by a subpart  of the
regulations or approved for a particular application.  A gauge  capable of
measuring  temperatures  to  within  3°C should be  provided  to monitor the
temperature around the filter during sampling.

     4.1.7   Sample transfer lines - A sample transfer line may be  used if
needed to  direct  sample flow  from  the  probe to the  filter or from the
filter to the  condenser.  The probe-to-filter line should be  insulated and
heated so  that  gas exit temperatures are  120°C  ±  14°C.  The  filter-to-
condenser  line  should  be  insulated  and  oriented with the downstream end
lower than the upstream end so  that any condensate will flow  away from the
filter and into the  condenser.   These  lines  should be  constructed of
Teflon®  or glass  and  should  be  recovered with their  respective  rinse
fractions  (front half or back half).

     4.1.8   Condenser - A multi-coil water-cooled  glass condenser should
be used to cool  the sample gas prior to entry  into the sorbent module.  The
orientation of the condenser should be vertical.

     4.1.9   Sorbent module - The  glass  water-cooled  container  configured
to hold the solid sorbent (XAD-2®)  should contain  a minimum of 20 g  of XAD-
2® and may contain  as much as 40 g.  A schematic diagram is shown in Figure
2.  A single piece  condenser-trap can be  used  if desired.  The sorbent trap
configuration  should   be  vertical  so  that  condensate  drains  from the
condenser  through the sorbent and  so that channeling  of  the  gas flow does
not occur.  The connecting fittings should form leak-free,  vacuum  tight
seals.  Sealant greases should  not be used in  the sampling train. A coarse
glass or Teflon® frit along with glass wool  plugs  is included to retain the
sorbent.    The  tester may  engrave  a  unique identification number for
inventory  and sample tracking.

     4.1.10   Impinger trains -  Four impingers should be connected in series
with  leak-free ground-glass  fittings  or  any  similar noncontaminating
fittings.  The first impinger should be a short  stem  (knock  out) version.
The second impinger should  be a Greenburg-Smith impinger with the standard
tip  and   plate.     The  third  and  fourth  impingers  should  be  the
Greenburg-Smith design modified so that the  glass tube has an  unconstricted
13  mm  ID  and  extends  to within  13  mm of  the flask  bottom.   The fourth
impinger outlet connection should  allow  insertion of a thermometer  capable
of measuring ±  1°C of  true value in the range of 0 to 25°C.

     4.1.11  Water circulating bath  - A bath and pump circulating system
which  is  capable  of  providing chilled  water flow to the  condenser and
sorbent trap water jackets  should be used.   Typically a submersible pump
is  placed  in the impinger  ice  water  bath so  that the  ice water contained
there  can  be  used.   The  function of this  system  should  be verified by
measuring  sorbent  trap gas entrance temperature
     4.1.12   Pitot  tube - The  pitot  tube,  preferably of Type S  design,
shall meet the  requirements  of  EPA  Method 2.  The  pitot  tube  is  attached
to the probe  as shown in Figure 1.  The proper pitot tube-sampling nozzle
configuration  for  prevention  of aerodynamic  interference  is  shown  in

                             0023A  -  4                         Revision 1
                                                             January 1995

-------
Figures 2.6 and 2.7 of EPA Method  2.  The Type S pilot tube assembly shall
have a known coefficient,  determined as outlined in Sec. 4 of EPA Method 2.

     4.1.13  Differential pressure gauge - The differential pressure gauge
should  be  an  inclined  manometer  or the equivalent  as described  in  EPA
Method 2.   Two  gauges  are  required:   one  gauge  to  monitor the  stack
velocity pressure  (AP),  and  the  other to measure  the orifice  pressure
differential (AH).

     4.1.14  Metering system - The metering system should consist of a dry
gas meter with 2% accuracy, a vacuum pump,  a  vacuum  gauge,  orifice meter,
thermometers or thermocouples capable of measuring  ± 3°C  of true value in
the  range  of 0  to 90°C;  and related  equipment  as shown  in Figure  1.
Thermocouples should be used  to monitor the  temperature  at  the following
sampling train locations:

     •  stack gas
     •  probe liner
     •  filter holder
     •  sorbent trap entrance
     •  silica gel impinger exit
     •  dry gas meter inlet and
     •  dry gas meter outlet.

Other metering  systems  capable of maintaining  isokinetic  sampling  rates
within  10%  and  determining  sample  volumes  to within  2%  may be  used  if
approved.   Sampling  trains with  metering  systems  designed  for  sampling
rates higher than those described in APTD-0581 and APTD-0576 (Air Pollution
Technical Document, see  references) may be used if the above specifications
can be met.  When the metering system is used  with a pitot tube, the system
should permit verification of an isokinetic sampling rate through the  use
of a nomograph or by calculation.

     4.1.15  Barometer  -  A  mercury  (Hg),  aneroid, or  other  barometer
capable of measuring atmospheric pressure to within ± 2.5 mm Hg is needed.
A  preliminary  check  of   a   new  barometer  should  be  made  against  a
mercury-in-glass barometer  or the  equivalent.   The absolute  barometric
pressure may be  obtained from a nearby weather service station and adjusted
for  elevation  difference  between the  station  and  the  sampling  point.
Either subtract 2.5 mm Hg from the station  value for every  30 m elevation
increase or add the same  for an  elevation  decrease.   If the  barometer
cannot be adjusted to agree within 0.1 in.  Hg of the reference barometric
pressure, it should be repaired or discarded.

     4.1.16  Gas density determination equipment - The equipment necessary
for conducting EPA Methods 2-4 for determining stack gas flow, molecular
weight  and  moisture  content,  respectively,   should  be  used.    Required
measurements  include  stack  gas   velocity  and  static   pressure;   gas
temperature; concentrations of 02, C02, and N2 (by difference), metered  gas
volumes and meter  temperatures  and  pressure; and condensate weight gain
collected by the impinger  train.   All equipment should meet EPA Method  2 -
4 requirements.
                             0023A - 5                         Revision  1
                                                             January  1995

-------
     4.2  Sample recovery equipment

          4.2.1   Fitting caps  - Ground glass  or  cleaned  aluminum  foil  to cap
     the exposed sections of the train.

          4.2.2   Wash bottles  - Teflon®, 500-mL.

          4.2.3   Probe-liner,  probe-nozzle, and filter-holder brushes - These
     should  be constructed  with  nylon  or  Teflon®  bristles  with  precleaned
     stainless steel or Teflon®  handles.  The probe brush should have extensions
     of stainless steel or Teflon® at least  as long  as the probe.   The brushes
     should be properly sized and shaped to  brush out the  nozzle,  probe liner,
     and front half filter holder.

          4.2.4   Filter storage container - Typically a glass petri dish sealed
     with Teflon® tape  is  used.  Petri dishes should  be  cleaned  according to
     glassware cleaning procedures listed in this method  (Sec. 6.1.4).

          4.2.5   Balance -  This balance  is  used  for measuring  weight gain of
     the impingers  and  sample  bottle weights  as  well.   Typically  a  0-2000-g
     balance is used.  The  balance  should  be accurate to within 0.5 g,  verified
     with Class S weights.

          4.2.6   Aluminum foil - Heavy duty cleaned by rinsing three times with
     methylene chloride and once with toluene, stored in pre-cleaned glass petri
     dish or glass jar.

          4.2.7   Graduated  cylinder  - Glass,  250-mL, with ± 1 ml resolution
     (this cylinder can be used for impinger volume  determinations in  place of
     the balance).

          4.2.8   Glass sample storage container - Amber glass bottle for sample
     glassware washes, 500- or 1000-mL, with leak-free  Teflon®-!ined caps.  The
     bottles should  be either purchased  as  precleaned or  cleaned  according to
     glassware cleaning procedures listed in this method  (Sec. 6.1.4)


5.0  REAGENTS

     5.1  Filters - Glass fiber filters, without organic binder, exhibiting at
least 99.95% efficiency (< 0.05% penetration) on 0.3 urn dioctyl phthalate smoke
particles.   All  filters  should be cleaned  before their  initial  use according to
the following procedures.

          5.1.1   Precleaning  - All  filters  should  be  cleaned before their
     initial  use.    Place  no  more than  50 filters  in  a Soxhlet  extraction
     apparatus.  Charge the Soxhlet with toluene and  reflux for 16 hours.  After
     extraction, allow the Soxhlet to  cool.   Remove the filters and dry under
     a clean  nitrogen (N2)  stream.   Store the filters in cleaned  glass petri
     dishes  or  amber glass bottles sealed with Teflon® tape  or Teflon®-! ined
     caps prior to  using them.
                                   0023A  -  6                         Revision 1
                                                                  January 1995

-------
           5.1.2    As  a  quality  control  check prior to the field test, take one
     precleaned filter and perform Soxhlet extraction with toluene for 16 hours.
     Remove  the toluene  extract and  analyze  according  to Method 8290.   No
     analytes may  be  observed above  the detection limit.

           5.1.3    Filter  surrogate  spike solution - As  stated  in  Sec.  7.3.3,
     Method 0023 calls  for  both  the  filter and the XAD-2® sorbent to be spiked
     with  the same set  of isotopically  labeled PCDD/PCDF standards.  Surrogate
     spikes  are added  to  the  sorbent  prior  to sampling  and  to  the  filter
     immediately before the sample extraction.  The filter and XAD-2® fractions
     (including the associated  glassware rinses) are extracted  separately and
     analyzed separately.   The  surrogate standards listed in Table 1 should be
     used  for both the  filter spike  and sorbent  spike.

           5.1.4    To  ensure proper  filter spiking, the isotopically-labeled
     standard solution, which is normally  at a concentration  of 0.1  ng//il_, is
     diluted to  0.004 ng//nL with nonane,  for  a  dilution factor of  25.   This
     spiking  solution  will  be   used  to spike the  surface of  the  filter as
     discussed in  Sec.  7.3.1.

     5.2   Sorbent  resin -  Amberlite XAD-2® resin.   XAD-2® may be  purchased
precleaned or cleaned by the laboratory.   If the resin  is not precleaned, the
cleaning procedures listed below shall be implemented.

           5.2.1    Sorbent  resin cleaning procedure

                   5.2.1.1   Place the sorbent resin in a  clean beaker and rinse
           with reagent water. Discard the rinse.  Fill  the beaker a second time
           with reagent  water and allow the resin to  stand overnight.  Discard
           this second rinse.

                   5.2.1.2   Place the sorbent  resin  in  an all-glass thimble of
           a  large  Soxhlet  extractor.   The sorbent  resin will  float when in
           contact with methylene chloride.  Therefore,  add a glass wool plug on
           top of the resin  in the thimble, and weight the glass wool plug down
           with a stainless  steel  ring  that fits  inside the  thimble.

                   5.2.1.3   Place the thimble filled with  resin into the Soxhlet
           extractor,  add  organic-free reagent  water to  the distilling  flask,
           apply heat, and extract the  resin for  8 hours.

                   5.2.1.4   Allow the  Soxhlet  extractor  to cool,  discard the
           water, and add methanol to the extractor.  Apply heat and extract for
           22 hours.

                   5.2.1.5   Again allowing  the  extractor  to  cool, drain off the
           methanol, replace it  with  methylene chloride.  Make sure  that the
           stainless  steel   ring  and  glass wool  plug are  still  in  place and
           extract  for 22 hours.

                   5.2.1.6   Extract  the resin  a  fourth time,  using  toluene as
           the extraction solvent, for  22 hours.
                                  0023A  - 7                         Revision 1
                                                                  January 1995

-------
                   5.2.1.7  Following the toluene extraction, the sorbent resin
          must be dried under a stream of clean dry nitrogen or other inert gas.
          This may be accomplished by transferring the resin to a large diameter
          glass  column  and  flowing the gas through the column.  The gas may be
          heated to less than 40"C, using a steam bath or other appropriate heat
          source.  Continue the  inert gas flow through the resin until all the
          residual solvent  is removed.   The  flow rate should be sufficient to
          agitate  the resin  particles,  but  not  so excessive as to  cause the
          particles to  fracture.

                   5.2.1.8  A quality control check should be conducted on the
          cleaned sorbent using HRGC/HRMS techniques  (Method  8290).  Typically,
          a  method  blank conducted previously on the same  lot  of  sorbent can
          serve  this  purpose.

          5.2.2    Sorbent resin surrogate spike  solution  - The-XAD-2® sorbent
      is spiked with isotopically  labeled PCDD/PCDF standards prior to sampling
      (surrogate  spikes).

      5.3  Glass  wool  - Cleaned  by  sequential  immersion  in  three  aliquots of
methylene chloride and one aliquot of toluene, dried in a 110°C oven, and stored
in a toluene-washed glass jar with a Teflon®-!ined screw cap.

      5.4  Organic-free  reagent  water  -   All  references to water in this method
refer to organic-free reagent water as defined in Chapter One.

      5.5  Silica gel  -  Indicating type,  6 to 16 mesh.  If previously used, dry
at 175°C for two  hours.  New silica gel may be used as received.  Alternatively,
other types  of desiccants  may be used, provided  that appropriate performance has
been demonstrated.

      5.6  Recovery solvents - Solvents must be  pesticide quality or equivalent.

          5.6.1    Acetone,  CH3COCH3

          5.6.2    Methylene chloride,  CH2C12

          5.6.3    Toluene,  C6H5CH3


6.0   SAMPLING COLLECTION, PRESERVATION,  AND  PREPARATION

      This section addresses preparation and collection procedures for sampling.

      6.1  Laboratory  preparation

          6.1.1    Filters.   (See Sec.  5.1.)

          6.1.2    Sorbent trap.  (See Sec.  5.2.)

          6.1.3    Glass wool - Precleaning and storage.   (See Sec.  5.3)
                                   0023A -  8                        Revision 1
                                                                  January 1995

-------
     6.1.4   Glassware  - All  glass  components  of the  train should  be
cleaned as follows:

     Soak all glassware  in hot soapy water (Alconox® or equivalent).
     Rinse with tap water to remove soap.
     Rinse with disti11ed/deionized H20 (three times).
     Bake at 400"C for 2 hours.
     Rinse with methylene chloride (pesticide grade) (three times).
     Rinse with toluene  (pesticide grade) (three times).
     Cap glassware with  clean glass caps or cleaned aluminum foil.
     Mark cleaned glassware with color-coded identification stickers.
     Rinse glassware immediately before  using with  acetone  and  methylene
     chloride.

     6.1.5   Because probe liners do  not  usually fit in glassware baths or
ovens, they may be rinsed three times with methylene chloride  followed by
three rinses with toluene, and sealed during transport.

6.2  Preliminary field determinations

     6.2.1   Sample  site -  The sampling  site  and  the minimum number  of
sampling  points should  be  selected  according  to  EPA Method  1  or  as
specified by the Agency.  The stack static pressure, temperature, and range
of velocity pressures (APs)  should be determined using  EPA  Method 2.   The
stack gas moisture content  should  be determined using  EPA  Method  4,  its
alternatives, previous data,  or an  engineering estimate.  Stack gas 02 and
C02 concentrations  should be  estimated and dry molecular weight  should be
calculated.   These  parameters  are used to estimate the isokinetic  sampling
rate settings.

     6.2.2   Nozzle size - The nozzle size should be based on the range of
velocity pressures so that it  is not necessary to  change the  nozzle  size
in order to maintain isokinetic sampling rates.

     6.2.3   Sampling duration - The total length of sampling  time  needed
to  obtain  the  identified minimum  sample gas  volume  is  determined  by
comparing  the  anticipated   average  sampling  rate  with   the   volume
requirement.   (Average  sampling  rate should  be within 0.5  to 0.75  cfm.)
The same time  should  be  allocated  to all traverse points defined  by  EPA
Method 1. To avoid timekeeping errors, the length of time sampled at each
traverse point should be an integer or an integer plus one-half minute.

             6.2.3.1  Calculation of length of the  sampling duration - The
     minimum sampling time required to  achieve a minimum sample volume and
     the corresponding detection limit (DL) are given below.

    ...  .         TO.                  analytical DL
    Minimum  sample  time  =
                          (Sample Rate)  x  (desired gas cone. DL)


             6.2.3.2  The  following  calculation is  for  a single  isomer
     (i.e., 2,3,7,8-TCDF).   Detection limits for other isomers may need  to
     be calculated as well.   For  this example, it will be  assumed that the


                             0023A -  9                         Revision  1
                                                             January 1995

-------
           analytical  detection limit  is  0.5 ng (actual  analytical  detection
           limit will  need  to  be specified  for each test program).

                   6.2.3.3   At a sampling  rate  of  0.014 m3/min  (0.5 cfm),  the
           sample  volume  per hour will  be  0.85 m3/h.   Assuming a desired stack
           gas concentration detection limit to be 0.1  ng/m3, the minimum sample
           time  required  to  collect  0.5 ng at  concentration  in the  stack of
           0.1 ng/m would  be:

             Minimum sample time =  	:	 = 6.25 h
                                   0.85 m3/h x 0.1  ng/m3


                   6.2.3.4   The total sampling  time should  be greater than or
           equal  to the  minimum  total  sampling  time  required  to  achieve  the
           necessary detection limit.  In addition, the sampling time per point
           should  be  greater than 2 min (greater minimum  time interval  may be
           specified by the Agency), and the sample volume  corrected to standard
           conditions  shall exceed the required minimum total gas sample volume.

     6.3   Calibration

     Calibration  of  the  apparatus is  one  of the most  important  functions in
maintaining data quality.  The detailed  calibration procedures for the sampling
apparatus listed in this section can be found in EPA Method 5 and Method 0010.
Table 4 summarizes the quality assurance functions for the calibrations.

           6.3.1    Metering system

                   6.3.1.1   Full  dry gas meter calibration - The dry gas meter
           (DGM)  in the meter  console  of  the sampling system should  be fully
           calibrated  against  a  primary  standard meter  (wet  test  meter or
           spirometer) or alternatively against a second reference meter (dry gas
           meter or critical orifice) that  has been calibrated  against a primary
           standard meter.   The procedure  can be found in Sees. 5.3 and 5.7 of
           EPA Method  5.

                   6.3.1.2   Post-test DGM calibration check - Following the  test
           program, the full  calibration factor  or meter Y should be checked by
           performing  a post-test DGM  calibration check. Any  secondary reference
           meters  can be used.   Three  calibration  runs  are  conducted  at  the
           maximum vacuum  reached  during  the testing.  The average post-test
           calibration factor should  not deviate from  the full DGM calibration
           factor  by  more than 5%.   Additional details on these procedures can
           be found in Sec. 5.3.2 of  EPA Method  5.

           6.3.2    Temperature gauges -  Each  thermocouple  should be permanently
     and   uniquely marked on  the  casting;  all  mercury-in-glass  reference
     thermometers should   conform  to  ASTM  E-l 63C  or  63F  specifications.
     Thermocouples should  be  calibrated in  the  laboratory with  and without the
     use  of extension leads.   If  extension leads  are used  in  the field,  the
     thermocouple readings at ambient air  temperatures,  with and  without the
                                  0023A - 10                        Revision 1
                                                                  January 1995

-------
     extension lead, should be noted and recorded.  Correction is necessary if
     the use of an extension lead produces a change greater than 1.5 percent.

                  6.3.2.1   Impinger,   organic  module,  and   dry   gas  meter
          thermocouples - For the thermocouples  used to measure the temperature
          of  the  gas  leaving  the  impinger  train and  the XAD-2®  resin  bed,
          three-point  calibration  at  ice-water,  room-air, and  boiling-water
          temperatures is necessary.  The thermocouples  should  be accepted only
          if the  readings  at all  three temperatures  agree  to  ± 2°C with those
          of the  absolute  value of  the  reference  thermometer.

                  6.3.2.2   Probe and stack thermocouple - For the thermocouples
          used  to indicate the  probe  and  stack  temperatures,  a  three-point
          calibration at ice-water,  boiling-water,  and hot-oil-bath temperatures
          should  be  performed;  it is  recommended  that  room-air temperature be
          added,  and that the thermometer and the thermocouple agree to within
          1.5%  at  each  of the  calibration  points.    A calibration  curve
          (equation) may be constructed  and the data extrapolated to cover the
          entire  temperature range  suggested by the manufacturer.

          6.3.3   Probe heater -  The probe heating system should be calibrated
     prior  to field  use  according to the  procedure  outlined  in  APTD-0576.
     Probes constructed  according to  APTD-0581  need not be calibrated if the
     curves of APTD-0576 are used.

          6.3.4   Barometer - The field  barometer should be adjusted initially
     and  before  each  test   series   to  agree  within   2.5  mm   Hg   of  the
     mercury-in-glass  barometer or  with  the station pressure value reported by
     a nearby National Weather Service station, corrected  for elevation.   The
     correction for  elevation difference between  the  station  and the  sampling
     point should be applied at a rate of -2.4 mm Hg/30 m of elevation increase.
     The results  should be  recorded on  the pretest sampling check form.

          6.3.5   Probe  nozzle  -  Probe nozzles  should be calibrated before
     initial use  in the field.   The ID of the nozzle should be measured with a
     micrometer to  the nearest 0.025  mm.   Three  measurements should  be  made
     using  different  diameters   each  time  and  the  average   obtained.    The
     difference between the high and the low numbers should not exceed 0.1 mm.
     When nozzles become  damaged they should not be used  again.   Each nozzle
     should be permanently  and uniquely  identified.

          6.3.6   Pitot  tube  -   The  Type  S pitot  tube  assembly  should  be
     calibrated using  the procedure outlined in EPA Method 2.

          6.3.7   Balance  - The balance should be  calibrated initially by using
     Class-S standard weights and should be within  0.5  g  of  the standard weight.

     6.4  Sampling train preparation  -  Care  should  be  taken to ensure a clean
sampling  train preparation  area free of excessive dust and organic compounds for
preparing the  sampling train.

          6.4.1   Preparation of impingers -  During preparation and assembly of
     the sampling train, all train openings where contamination  can enter should
     be sealed until  just  prior to assembly or until  sampling is about to begin.

                                  0023A - 11                         Revision 1
                                                                  January 1995

-------
                  6.4.1.1   The  first  impinger  should  be  left  empty (used  as  a
          water knock-out impinger due to long run times).

                  6.4.1.2   Approximately 100 ml  of  reagent  water should  be
          placed  in the  second  and third  impingers.    This  method does  not
          require that organic analyses be conducted  on the impinger contents.
          However,  if analyses of  semivolatile  organic  compounds are to  be
          conducted,  then the proper specifications on cleaning  the impingers
          and water quality  (i.e., HPLC-grade water)  should be observed.

                  6.4.1.3   Approximately 200 to  300 g  of silica  gel should be
          placed  in  the  fourth  impinger.   All   impingers  should  be  weighed
          separately  to the nearest 0.5 g and the weights recorded.  Impingers
          should be connected with glass U-tube connectors.

          6.4.2   Filter   loading  -  A  filter  should   be  placed  in   a
     properly-cleaned filter holder using cleaned tweezers or clean disposable
     surgical gloves.   The filter should be properly centered and the gasket (if
     used) properly placed to prevent the sample gas  stream from  circumventing
     the filter.  The filter should be checked for tears  after the assembly is
     completed.

          6.4.3   Sorbent loading  - The  XAD-2®  should  be loaded  and sealed in
     the analytical (preparation) laboratory.

          6.4.4   Final  assembly  - The  final  assembly of  the filter  holder,
     condenser, and sorbent module can  be performed at  the stack location.   All
     components should be sealed with either precleaned foil or socket  joints.

     6.5  Sampling  train  leak check procedures  -  Leak  checks  are necessary to
assure that the sample has not been biased low  by dilution air.   Both pre-test
and post-test leak checks are necessary.

          6.5.1   Pre-test  - After  the sampling  train  has been  assembled,  the
     train should be  leak checked  at the sampling site by plugging the nozzle
     and  pulling  a 380 mm  Hg  vacuum.    Leakage  rates  greater than 4%  of  the
     average   sampling   rate  or  0.00057 m3/min,  whichever  is   less,   are
     unacceptable.  Leak  checks should be  conducted according to EPA Method  5
     criteria.

          6.5.2   During  the sampling  -  If a component (e.g.,  filter assembly,
     sorbent module,  or impinger) change is necessary during the  sampling  run,
     a leak check should be conducted  before the  change.  The leak check should
     be done according  to the  procedure  outlined above,  except that it should
     be at a vacuum equal to or greater  than the maximum value  recorded up to
     that point in  the test.  If the leakage is less than 0.00057 m3/min or 4%
     of  the  average   sampling  rate   (whichever  is   less),  the  results  are
     acceptable.   If, however,  a higher leakage  rate  is obtained,  the tester
     should record  the leakage rate and either  void the sampling run or perform
     sample  volume leak corrections  (if  approved  by  the  Agency).    After
     replacing  the  train  component,  an initial  leak check should be completed
     before sampling.
                                  0023A - 12                        Revision 1
                                                                  January 1995

-------
     6.5.3   Post-test  -  The  leak check should be completed at  a  vacuum
equal to or greater than the maximum value reached during the sampling run.
If  the  leakage rate  is  less  than 0.00057  m3/nrin  or 4%  of the  average
sampling  rate  (whichever  is  less),  the  results  are  acceptable.    If,
however, a higher  leakage  rate is obtained the tester shall either void the
sample run or perform sample volume leak corrections  (if  approved  by the
Agency).

6.6  Sampling train operation

     6.6.1   Final pre-test sampling checks  - After conducting the initial
leak check, the following checks should be  made:

  •  Meter box  examination
  •  Manometers  leveled  and zeroed
  •  Pump checked  for proper operation
  •  Pitot lines  leak checked
  •  Probe markings  verified
  •  Thermocouples reading correctly
  •  Size and orientation  of the nozzle verified
  •  EPA Method 3  equipment  for C02/02  checked for proper assembly and leak
     checked  and
  •  Isokinetic  K-factor checked to ensure that it  is  correct.

Immediately prior to sampling:

  •  Portholes  should  be  cleaned to  minimize  the  chance  of  sampling
     deposited material
  •  Probe and  filter heating  system temperatures  should be checked
  •  Condenser/sorbent cooling  system  temperatures  should  be checked and
  •  Proper nozzle location should be  verified.

     6.6.2   The  sampling procedure below should be followed.

             6.6.2.1  Sampling   -  Initial   dry   gas  meter   readings,
     barometric pressure,  and  temperatures should be recorded.  The tip of
     the probe should be  positioned at the  first sampling  point  with the
     nozzle tip pointing directly into the gas  stream.   When the probe is
     in position,  the open  area around the  probe and  the  porthole  should
     be blocked off  to  prevent flow disturbances and non-representative
     dilution of  the  gas  stream.   The pump should  be turned on and  the
     sample flow adjusted  immediately to attain  isokinetic conditions.  The
     EPA Method 3  sampling system  should be turned on.  Velocity pressures
     should  be  recorded  and  the  sampling   rate  adjusted  to  isokinetic.
     Other readings of velocity pressure (AP), orifice pressure (AH), stack
     gas temperature (TJ, probe temperature  (Tp), filter temperature (Tf),
     sorbent  trap temperature (Tt), silica gel   impinger temperature  (Tsg),
     dry  gas  meter  inlet and outlet  temperatures  (Tm),  dry  gas  meter
     volume,  and sample  vacuum should  be made.

             6.6.2.2  The stack should be  traversed  as directed  in  EPA
     Method 5 procedures.   At each  sample point, the above readings  should
     be taken and  sample flow rates adjusted to  isokinetic.  Following the


                            0023A - 13                         Revision 1
                                                             January 1995

-------
          traverses, the pump is turned off, the probe removed from the stack,
          and the final DGM readings recorded.  Care should be  taken not to bump
          the  nozzle  against stack walls  in  order to minimize the  chance  of
          breakage  or  extracting  deposited  material.    Following  each  port
          traverse, a leak check is recommended  in  order to ensure a leak tight
          system.   An additional leak  check  may also be  performed  after the
          train  is  moved  to the next port, prior  to  sampling.   The  necessary
          post-test leak check should be conducted and the leak rate  recorded.

                  6.6.2.3   Periodically during  the  test  run,  the  connecting
          glassware from  the probe,  through the filter,  and  to the  condenser
          should  be checked for water  condensation.   If  any  condensation  is
          evident,  verify  that  the  temperature  sensors  and heater systems are
          functioning properly.  Ice should be maintained around the  impingers
          to keep both the  sorbent trap entrance and silica gel exit temperature
          at 20°C.  Filter vacuum should be checked for sudden increases.   The
          filter  should be  changed if  the  vacuum  exceeds  15 in.  Hg.    The
          manometer level  and zero  should  also  be  checked periodically during
          each  traverse,  because vibrations  and temperature  fluctuations can
          cause the manometer zero to shift.

                  6.6.2.4   Following the post-test leak check,  the probe should
          be disconnected, and the nozzle and the end of the probe capped with
          precleaned aluminum foil,  or equivalent caps.  The inlet to the filter
          holder  should be capped  according to one of  the  methods  previously
          mentioned.  It may be necessary to loosen the seal between the sorbent
          module  outlet and the inlet  to the first impinger  to prevent  water
          from  being  drawn back into the module when the sample  train cools.
          Alternatively, the filter holder, condenser  and  sorbent module may be
          disassembled  and immediately capped at the stack location and removed
          to the  sample recovery area.

     6.7  Collection  of blanks  -   Four  different sampling  blanks  should  be
collected:   field  blanks,  reagent blanks,  proof blanks,  and method  blanks
(laboratory only).  Only two sampling blanks should be analyzed initially:  the
field blank and the  laboratory method blank.  If the field  blank has high levels
of contamination and the laboratory blank does not show high background levels
of PCDD/PCDF, the other blanks should be analyzed to help determine  the source
of the contamination.   Blanks are further discussed in Sec. 8.0.
7.0  PROCEDURE

     7.1   Recovery preparation - Proper recovery procedure begins as soon as the
probe is removed from the stack at the end of the sampling period.   The nozzle
end of  the  sampling  probe should be sealed with  precleaned  aluminum foil  and
disconnected from the filter holder.  When the probe is cool enough to be handled
safely, all  external  particulate matter near the tip of the probe  should be wiped
off and both ends of  the probe  closed off with aluminum foil.  Both openings to
the filter holder, transfer line (if used), condenser,  sorbent trap, and impinger
train should be disconnected and sealed.  Care should be taken not to lose any
condensed water upstream of the impingers (if present) during this process.
                                  0023A - 14                        Revision 1
                                                                  January 1995

-------
     Train  components  should be transferred  to  the cleanup area.   This  area
should be clean and enclosed so  that the chances of losing or contaminating the
sample are  minimized.   Smoking, which  could contaminate  the  sample,  is  not
allowed in the cleanup area.  Cleanup personnel should wash their hands prior to
sample recovery.  The train  should  be  inspected prior to and during disassembly
and any abnormal conditions, e.g., broken filter, colored impinger liquid, etc.,
noted.

     7.2   Sample  recovery procedure - As shown in Figure 3,  the sampling train
should be  recovered  into  four containers.   The procedures  applicable  to  each
sample container are briefly discussed in the following section.

           7.2.1    Filter  (Container 1) - The filter  should be removed carefully
     from  the filter holder and placed in  its identified container.   Cleaned
     tweezers  should be  used  to  handle  the filter.    Fold  the filter,  if
     necessary, with the  particulate cake  inside  the  fold.   Any particulate
     matter and filter  fibers which adhere  to the  filter  holder  gasket should
     be transferred  to  the  container  by  using a dry inert  bristle brush and a
     sharp-edged  blade.  The container should be sealed with Teflon® tape.

           7.2.2    Front half rinse (Container 2)  - The front half  glassware
     surfaces  will  be  rinsed  with  acetone  and  brushed,  followed  by  three
     additional rinses  with methylene chloride.  All rinses should be put  into
     Container  2.   The  outside of the  probe,  the  pitot tube,  and the nozzle
     should be cleaned to  prevent particulates  from  being brushed  into  the
     sample  bottle.   The  probe  liner  should be  tilted  and rotated  while
     squirting  acetone  into the upper end  to assure complete wetting  of  the
     inside surface. Acetone is then  squirted into the upper end  while pushing
     the probe brush through the liner with a twisting motion, with the drainage
     caught in the sample bottle (Container 2).  The brushing procedure should
     be  repeated   two more  times  or  until  no  particles  are  visible  in  the
     drainage  and  a visual  inspection  of the  liner reveals  no  particles
     remaining  inside.  The brush  should be rinsed  into the sample  bottle to
     collect  any  particulates  that may be retained within  the bristles.   The
     three  rinses should  be repeated  with methylene  chloride  allowing  the
     rinsate  to collect into the same sample container.

           After all  the rinsings  have been  collected,  the  lid on the sample
     container  should   be  tightened   securely.   As  a  precaution  in  case  of
     leakage, the liquid level  should  be marked on the sample container and the
     cap sealed with Teflon® tape.  The  sample recovery should be recorded on
     the sample recovery form.

           7.2.3    Sorbent module (Container  3) - The sorbent module  should be
     removed  from  the train, tightly capped at both ends with aluminum foil or
     glass  caps,  labeled  and stored  on  ice for  transport  to  the  laboratory.
     Care  should be taken to ensure that no ice water can  leak  into the stored
     traps or any  other train component.

           7.2.4    Back  half rinse  (Container  4) - The  back  half  of the filter
     holder,  transfer  line  (if used)  and condenser  (if  separate from  trap)
     should be  rinsed three times  with acetone followed by three  rinses  with
     methylene chloride.  The sample container (Container 4) is  then identified
     and sealed as discussed above.

                                 0023A -  15                       Revision 1
                                                                  January  1995

-------
          7.2.5    Impinger  water - Any  color or  film  in the  impinger  water
     should be noted on the  sample recovery form.   The  entrained moisture in the
     first  three  impingers  should  be measured  to within ±  1  mL  by  using  a
     graduated cylinder or  by weighing to within 0.5 g by using a balance, and
     the data recorded appropriately.  This information is needed to calculate
     the moisture  content of the effluent gas.

          7.2.6    Silica gel - The color of the indicating silica gel should be
     noted on the recovery form to determine if it has  been completely spent and
     the impinger weighed to determine entrained moisture weight  gain.  Analysis
     is not required.

     7.3  Analysis summary  -  The following section summarizes  the analytical
procedures for quantitating  PCDD/PCDF collected by the sampling train.   Sample
preparation procedures  and  the basic analytical  techniques  are listed.   The
detailed analytical protocol can be found in Method 8290.

          7.3.1    As  shown  in  Figure  4,  the analytical  procedure requires the
     sampling train to be analyzed  in two fractions. Containers  1 and 2 (filter
     and  front   half  rinse) are combined  and analyzed.   Containers 3  and  4
     (sorbent trap and back  half rinse) are also combined and analyzed.  In this
     way  filter  surrogate  standard  recoveries  and XAD-2® surrogate standard
     recoveries  are both determined separately.

          7.3.2    Acceptance  criteria  and corrective  actions   for  surrogate
     recoveries  are as  follows:

                   7.3.2.1   All  PCDD/PCDF surrogate recoveries should be within
          70 to  130 percent.

                   7.3.2.2   If   all   isomer  recoveries   are   greater   than
          130 percent,  the  sampling runs  should  be repeated,

                   7.3.2.3   If all  isomer recoveries are less  than 70 percent,
          the  sampling runs  should  either be repeated or the  final  results
          should be divided by  the fraction of surrogate  recovery.

                   7.3.2.4   Acceptance criteria  for other standard recoveries
           (i.e., internal)  should  conform to  Method 8290  requirements.

          7.3.3    As  discussed  in  Sees.  5.1.2 and 5.2.1, surrogate spikes are
     added  to the  sorbent trap prior to sampling and to the filter immediately
     prior  to  extraction.   The  same set  of isotopically-labeled compounds is
     used  for  these  spikes.   The  analytical  procedure for both fractions is
     given  in the  following sections.   All  samples should be  extracted within
     30 days of  collection  and  analyzed  within 45  days  of extraction.

          7.3.4    Sample  preparation  and  internal  standard  addition  -  The
     following procedure should be performed for  the filter/front half analysis
     and  the sorbent  trap/back half analysis.  The only difference between the
     two  procedures  is that surrogate  standards  are added to  the filter/front
     half fraction immediately prior to sample preparation whereas the surrogate
     standards have  already been added  to the sorbent trap/back half prior to
     sampling.

                                  0023A - 16                        Revision  1
                                                                  January  1995

-------
        7.3.4.1   Filter/front  half  fraction  procedures  -  Place  a
cellulose extraction thimble, 1 g of silica gel or sodium sulfate,  and
a plug of glass wool  into the Soxhlet apparatus,  charge the apparatus
with toluene, and reflux for a minimum of 3  hours.  Remove the toluene
and discard  it,  but retain the  silica  gel.   Remove  the  extraction
thimble from the extraction system and place it  in  a glass beaker to
catch the solvent rinses.

        7.3.4.2   Add exactly 1.0 ml of the surrogate spiking solution
(Sec. 5.1.2) uniformly onto  the surface of the  filter while  it  is
still in the petri dish in which it  was returned from the field,  using
an adjustable pipet.  Transfer the  filter  directly  to the  extraction
thimble of the extraction  system.   Rinse the petri dish with 10  ml of
toluene three times collecting the rinsate into  the beaker.

        7.3.4.3   Concentrate  the  sample   in  Container 2  (acetone/
methylene  chloride   rinses)  to a  volume  of  about  1-2 ml_  using  a
Kuderna-Danish concentrator apparatus, followed by nitrogen blow down
at a temperature of  less than 37°C.  Rinse the  sample container  three
times with small portions  of  methylene chloride  and add these to  the
concentrated solution and  concentrate further  to  near dryness.  This
residue contains particulate  matter removed in the rinse of the  train
probe and nozzle.  Add the concentrate  to the filter  in the Soxhlet
apparatus described above.

        7.3.4.4   Add 40 juL   of  the  internal   standard   solution.
Fortification  is accomplished  by  using  the  sample  fortification
solutions described in Table  1.  Cover the contents  of the  extraction
thimble with the cleaned glass wool  plug  and proceed to the  extraction
procedure.

        7.3.4.5   Sorbent trap/back half  fraction procedures - Prepare
another  extraction  thimble/silica  gel  system as  described above.
Suspend the adsorbent module  directly over the extraction  thimble in
the  beaker.    The glass  frit  of  the module  should  be  in the  up
position.  Using  a Teflon® squeeze  bottle containing  toluene,  flush
the  XAD-2®  into  the  thimble onto  the  bed of  cleaned silica gel.
Thoroughly rinse the glass module catching the rinsings in  the beaker
containing the  thimble,  first with  methanol, if needed,  then with
toluene into the thimble.   If the resin  is  wet, effective  extraction
can be accomplished  by loosely  packing the resin  in  the thimble.  Add
glass wool plug from the XAD-2® sampling module  to  the thimble.

        7.3.4.6   Concentrate    the    sample    in    Container  4
(acetone/methylene chloride  rinses)  to  a  volume of  about 1 -  2  ml
using a Kuderna-Danish concentrator  apparatus, followed by  nitrogen
evaporation at  a  less than 37°C.   Rinse the sample container  three
times with small portions  of  methylene chloride  and add these to  the
concentrated solution and concentrate further to near  dryness.   Add
the concentrate to the XAD-2®  resin  in the Soxhlet apparatus described
above.

        7.3.4.7   Add  40  /nL  of  the  internal   standard   solution.
Fortification  is accomplished  by   using  the sample  fortification

                       0023A  - 17                         Revision  1
                                                       January 1995

-------
          solutions described in Table 1.   Cover the contents of the extraction
          thimble  with  a  cleaned glass wool plug to prevent  the  XAD-2® resin
          from floating into the solvent reservoir of the extractor and proceed
          with extraction  (Sec.  7.3.2).

          7.3.5    Sample extraction - Place the thimble in the extractor and add
     the  toluene  contained  in  the  beaker to  the  solvent  reservoir.    Pour
     additional  toluene  to fill the reservoir approximately  two-thirds full.
     Add  Teflon® boiling  chips  and  assemble  the  apparatus.   Adjust  the  heat
     source to cause the extractor to cycle three times per hour.   Extract the
     sample for 16 hours.   After extraction, allow the Soxhlet to cool. Transfer
     the  toluene  extract  and  three  10-mL  between  rinses  to  the  rotary
     evaporator.   Concentrate the extract  to approximately 10 ml.

          Use a  nitrogen  evaporative concentrator to reduce  the  volume  of the
     extract to  about 100  juL.   Redissolve  the residue  in  5 mL of  hexane.

          7.3.6    Sample clean-up and fractionation  - Sample extracts described
     above  are  spiked  with  40 pi  of  the  alternate  standard  fortification
     solution, then divided  into two equal  portions.   One half  of each  sample
     extract  is  archived  for future needs.   The  other portion  is  solvent-
     exchanged to hexane then subjected  to  three column chromatographic cleanup
     steps as described in Method 8290.

          7.3.7    Analysis summary  -  The  samples  are  analyzed  with  a  high
     resolution  gas  chromatographic  column coupled  to a  high resolution  mass
     spectrometer  (HRGC/HRMS) using the instrumental  parameters described below.
     Prior to analysis,  the Recovery Standard  solution  from Table  1 is added to
     each sample.  Sample  extracts are first analyzed using a capillary column
     to determine  the concentration  of  each isomer  of  PCDDs  and  PCDFs (tetra-
     through  octa-).   If  2,3,7,8-TCDF  is  detected  in this  analysis,  another
     aliquot of  the  sample is analyzed  separately,  using a second, dissimilar
     column  to  confirm and  more accurately measure the  2,3,7,8-TCDF isomer.
     Other  column systems  may  be  used,  provided  that  the  user is able  to
     demonstrate by means of calibration and performance checks that the column
     system  is able to meet  the specifications of Sec. 5.6 of Method 8290.

          7.3.8    All  other  analytical  specifications   for  determining  the
     amounts of PCDD/PCDF isomers collected  in the filter/front half and sorbent
     trap/back half fractions can be  found  in Method 8290.

     7.4  Calculations

     The  following section describes the  calculations used to determine gas
concentrations  and emissions  of PCDD  and  PCDF  isomers.   Toxic  equivalent
calculations are not included in this method.   Each  set of calculations  should
be repeated  or spot-checked, as a QC measure.  Calculations should be carried out
to at least  one extra decimal  place beyond  that of the acquired data and should
be rounded off after final  calculation to two significant  digits for each run or
sample.   All  rounding of  numbers  should  be performed  in  accordance  with the
ASTM 380-76 procedures.

     The  nomenclature and  sampling equations are presented in Sec. 7.4.1.


                                  0023A -  18                        Revision 1
                                                                  January 1995

-------
 7.4.1    Sampling nomenclature

An     =  Cross  sectional area of nozzle, m2 (ft2).

As      = Cross  sectional area of stack, m2 (ft2).

BW8     = Water  vapor  in the  gas stream, proportion  by  volume.

C;       = Concentration of pollutant  i, jug/dscm (Ib/dscf).

E|       = Emission rate of pollutant  i, g/sec  (Ib/hr).

DN      = Diameter of  nozzle, mm (in.)

I       = Percent of isokinetic sampling.

Mw      = Molecular weight of water, 18.0 g/g-mole  (18.0 Ib/lb-mole).

Md      = Molecular weight of dry stack gas, g/g-mole  (Ib/lb-mole).

M8      = Molecular weight of wet stack gas, g/g-mole  (Ib/lb-mole).

ttij       = Mass of pollutant  i collected by sampling  train,  /ng (lb).

Pbar     = Barometric pressure at the  sampling  site,  mm  Hg  (in.  Hg).

Static    = Static gauge pressure of stack gas,  mm H20 (in. H20).

P8      = Absolute stack gas  pressure, mm Hg (in.  Hg).

Pstd     = Standard absolute pressure, 760 mm Hg  (29.92  in.  Hg).

Qsd      = Average stack gas volumetric flow, dry, standard conditions,
          dscmm  (dscfm).

R       = Ideal gas constant,  0.06236  [(mm Hg)  (m3)] / [(°K)  (g-mole)]
          (21.85 [(in. Hg) (ft3)]  / [(°R)  (lb-mole)]}.

Tm      = Absolute average DGM temperature°K (°R).

Ts      = Absolute average stack gas  temperature'K (°R).

Tstd     = Standard absolute temperature, 293°K (528°R).

V,c      = Total  volume liquid collected  in  impingers and silica gel
          (ml).

Vm      = Volume of  gas  sample as  measured by dry gas  meter, dcm
          (dcf).

       =  Volume  of  gas  sample  measured  by  the  dry gas  meter,
          corrected to standard conditions, dscm  (dscf).


                        0023A -  19                        Revision 1
                                                        January 1995

-------
     Vw
-------
     7.4.4   Moisture content

                                  V
                        B   =
                                   w(std)
                              V     + V
                               m(std)   w(std)

NOTE:  In saturated or water droplet-laden gas streams,  two calculations of
       the moisture content of the stack gas should be made, one from the
       impinger  analysis  (Sec. 7.4.3), and a  second from the assumption of
       saturated conditions.   The  lower  of the two  values of Bws should be
       considered correct.   The  procedure  for determining  the moisture
       content based  upon assumption of  saturated  conditions is given in
       the  Note   in Sec.  1.2  of EPA Method 4.   For the purposes  of this
       method, the average stack gas temperature may be used to make this
       determination,   provided   that   the   accuracy  of   the  in-stack
       temperature sensor is  ± 2°C.

     7.4.5   Absolute stack  gas pressure

                                     P   .
                          p  _ p   +  static
                               bar    13.6


     7.4.6   Average molecular weight of dry stack gas

Dry: Md = (0.32  x %02) x  (0.44 x %C02) +(0.28 x (100 -  (%02 + %C02))


                Wet:  M -  M. x (1  - B  )  + (B   x M  )
                      s    d   v      ws'   v ws   w'


     7.4.7   Stack gas velocity at  stack conditions
                  Vs = K x C  x
                                           * T
                                              std
                                         P x M
     7.4.8   Average stack gas volumetric flow at dry, standard conditions

                                        d  X P*   60sec
            Qd = V  xA  x (1  -B   ) x
            ^sd    s    s   x     ws '
                                      T x P  .     min
                                       s    std
     7.4.9   Concentration of pollutant

                                   M.
                             C. = —'-
                             0023A -  21                         Revision 1
                                                             January 1995

-------
          7.4.10   Emission of pollutant
           7.4.11   Isokinetic  sampling rate

                            1039.5746 x V   x  (T  + 460)
                           V, X0x Ps x  (1  -BwJ  x (DJ2
                                *English units
8.0  QUALITY CONTROL

     The  following  quality  control  (QC)  guidelines  outline pertinent steps to
be followed during  the production of emission  data  to  ensure  and  quantify the
acceptability and reliability of the data generated.

     8.1   Sampling  QC procedures - Quality control procedures specific to manual
source gas sampling  procedures should follow EPA Method 5 and those  listed in EPA
Manual  600/4-77-0276 for  Method  5.   Sampling QC procedures are  summarized in
Table 2.

     8.2   Blanks

           8.2.1    Field blank - A field blank should be  collected  from a set of
     glassware that has not been used to collect any field samples.  The field
     blank train is loaded,  leak checked, and left at a sampling location during
     a  test run.  The train is then recovered.  The purpose of the field blank
     is  to measure the  level  of  contamination that  occurs from  handling,
     loading,  recovering, and transporting  the  sampling train.    Collect one
     field blank for every  nine  test runs  at each test  location.

           8.2.2    Glassware blank (proof blank) -  A proof blank  is  recovered
     from each set of  sampling train glassware that  is used to collect the
     organic  samples.   The precleaned glassware,  which consists of  a  probe
     liner, filter  holder, condenser coil,  and  impinger set, is loaded as if for
     sampling  and then quantitatively recovered exactly as  the samples will be.
     Analysis  of  the  generated  fractions  will be  performed  to  check  the
     effectiveness  of the glassware cleaning procedure,  but only if field blank
     analysis  indicates a potential contamination problem.   If requested by the
     test administrator,  collect one glassware blank per each set of glassware
     used on  the  complianca test and archive for future analysis  in  the event
     the  field blank shows  contamination.

           8.2.3    Reagent blank -  Reagent  blanks should  contain 500 mL of each
     reagent  used  at the test  site.   Reagent blanks are  saved  for  potential
     analysis.   Each reagent  blank is part of  the same  lot  used during the
     sampling  program.     If  a  field  blank   is   unsatisfactory because  of
     contamination, reagent blanks may be  analyzed to  determine  the specific

                                  0023A - 22                        Revision  1
                                                                  January 1995

-------
     source of contamination.  Collect one reagent blank per compliance test and
     archive  for  future  analysis  in  the event  that  the field  blank  shows
     contamination.

          8.2.4    Laboratory method blank -  A  method  blank  is  a performance
     control sample that is prepared in the laboratory and processed in a manner
     identical  to  a field  sample.   The  XAD-2®  resin should  be  from the same
     batch used for preparation of the field traps.  One  laboratory method blank
     should be  analyzed for every batch  of samples  analyzed.


9.0  METHOD PERFORMANCE

     9.1  Method performance evaluation  - Evaluation of analytical procedures
for  a   selected  series  of  compounds  shall  include  the sample  preparation
procedures  and   each  associated  analytical   determination.    The  analytical
procedures should  be  challenged  by  the  test compounds  spiked  at appropriate
levels  and carried through the procedures.

     9.2  Method detection  limit  -  The  overall  method detection limits (lower
and upper) should be calculated as shown in Sec.  6.2.3.1.  Generally, analytical
detection limit  for tetra-CDD/CDF congeners are 50  pg.  Penta-,  hexa-,  and hepta-
congener  detection  limits are 250  pg  and octa-congener  detection  limits  are
500 pg.

     9.3  Method  precision and bias -  The overall  method precision and bias
should  be determined on a compound-by-compound  basis  at a given concentration
level.    The method precision value  includes  a  combined variability due  to
sampling, sample preparation,  and instrumental  analysis.  The  method  bias  is
dependent upon the  collection, retention, and extraction efficiency of the train
components.  Interlaboratory testing of Method 0023 and Method  8290 to establish
method  accuracy and precision for sampling a variety of stationary sources has
not been performed.


10.0 REFERENCES

1.   American Society of  Mechanical Engineers,  Sampling for  the Determination
     of Chlorinated Organic Compounds  in Stack Emissions.   Prepared for U.S.
     Department of Energy and U.S. Environmental  Protection Agency.  Washington,
     DC.  December 1984.

2.   American Society of  Mechanical  Engineers.  Analytical Procedures to Assay
     Stack Effluent Samples and Residual Combustion Products for Polychlorinated
     Dibenzo-p-Dioxins   (PCDD)    and  Polychlorinated   Dibenzofurans  (PCDF).
     Prepared  for  the  U.S.  Department  of Energy  and  U.S.  Environmental
     Protection  Agency.   Washington, DC.   December  1984.

3.   Thompson, J.R.,  Analysis of Pesticide Residues in  Human  and Environmental
     Samples,  U.S. Environmental  Protection Agency, Research Triangle Park, NC,
     1974.

4.   U.S.  Environmental  Protection Agency.    Method 8290:   The Analysis  of
     Polychlorinated  Dibenzo-p-Dioxins  and  Polychlorinated   Dibenzofurans  by

                                  0023A - 23                        Revision 1
                                                                  January 1995

-------
     High-Resolution Gas Chromatography/High-Resolution Mass Spectrometry. In:
     Test Methods  for  Evaluating  Solid Waste,  SW-846.  Washington,  DC.

5.   U.S.  Environmental  Protection  Agency.    Method  8280:  The Analysis  of
     Polychlorinated Dibenzo-p-Dioxins  and Polychlorinated Dibenzofurans. In:
     Test Methods  for  Evaluating  Solid Waste,  SW-846.  Washington,  DC.

6.   Tondeur,  Y.,  Albro,  P.W.,  Mass,  R.J.,  Harvan,  D.J.,  Schroeder,  J.L.,
     "Matrix Effect in  Determination of 2,3,7,8-Tetrachlorodibenzodioxin by Mass
     Spectrometry", Anal.  Chem. 56(8), pp  1344-1347, 1984.

7.   Tondeur,  Y.,  Niederhut,  W.N.,  Campana,  J.E.,  Missler,  S.R.,  "A Hybrid
     HRGC/MS/MS  Method for the Characterization of Tetrachlorinated-p-Dioxins
     in Environmental  Samples", Biomed. Environ. Mass Spectrom. 14(8), pp 449-
     456, 1987.

8.   Taylor, J.K., Quality Assurance of Chemical Measurements,  Lewis Publishers,
     Inc.,  1987.

9.   Department of Health, Education, and Welfare, Public Health Service, Center
     for Disease Control.  Carcinogens - Working with Carcinogens.  Publication
     No. 77-206.  National Institute for Occupational Safety and Health.  August
     1977.

10.  OSHA  Safety  and  Health Standards, General  Industry.   29  CFR,  p   1910.
     Occupational  Safety and Health Administration.  OSHA 2206. Revised January
     1976.

11.  American  Chemical  Society,   Committee  on  Chemical  Safety.    Safety  in
     Academic  Chemistry Laboratories.  3rd Edition,  1979.

12.  40 CFR Part 60, Appendix  A.

13.  Martin, R.M., Construction Details of Isokinetic Source-Sampling Equipment.
     U. S.  Environmental Protection Agency, Research Triangle Park,  NC.   Air
     Pollution Technical  Document (APTD) 0581,  April 1971.

14.  Rom, J.J.,   Maintenance,  Calibration, and Operation of  Isokinetic Source
     Sampling  Equipment.    U.S.   Environmental  Protection  Agency,  Research
     Triangle  Park,  NC.  Air  Pollution Technical  Document (APTD)  0576,  March
     1972.
                                  0023A - 24                        Revision 1
                                                                  January 1995

-------
                                    TABLE  1

                   COMPOSITION OF THE SAMPLE FORTIFICATION
                       AND RECOVERY STANDARDS SOLUTIONS
Analyte
Internal Standards
13C12-2,3,7,8-TCDD
13C12-l,2,3,7,8-PeCDD
13C12-l,2,3,6,7,8-HxCDD
13C12-l,2,3,4,6,7,8-HpCDD
13C12-OCDD
13C12-2,3,7,8-TCDF
13C12-l,2,3,7,8-PeCDF

13C12-l,2,3,6,7,8-HxCDF

13C12-l,2,3,4,6,7,8-HpCDF
Surrogate Standards

37Cl4-2,3,7,8-TCDD

13C12-l,2,3,4,7,8-HxCDD
13C12-2,3,4,7,8-PeCDF
13C12-l,2,3,4,7,8-HxCDF
13C12-l,2,3,4,7,8,9-HpCDF
Recovery Standards
13C12-1,2,3,4-TCDD
13C12-l,2,3,7,8,9-HxCDD
Alternate Standard
13C12-l,2,3,7,8,9-HxCDF
0023
Concentration
(P9/ML)

100
100
100
100
100
100
100
100

100


100

100
100
100
100

500
500

100

82901
Concentration
(P9/ML)

10
10
25
25
50
10
10


25






25


50
50

_ _

1  Provided  as  reference only;  also see  Tables  2  and  3  of Method  8290.
                                  0023A - 25
  Revision 1
January 1995

-------
           TABLE 2

SAMPLING QC PROCEDURES SUMMARY
QC Procedure
Sample equipment
calibrations
Dry gas meter sample
leak check
02 and C02 sampling
system leak check
AP meter leveling
Pitot tube leak check
Pitot tube orientation
check
Cyclonic flow check
Probe, filter, trap,
and silica gel impinger
are maintained at
specified temperature
ranges
Overall isokinetic
sampling rate
Sampling blanks
Frequency
See Sec. 6.3.1
Before and after
each test run
Once per test
Before and after
each test run
Before and after
each test run
Every test
Made at every
location
Every test
Every test
See Sec. 8.2
Criteria
See Sec. 6.3.1
0.00057 cmm (< 0.02
cfm) or 4% of sample
rate whichever is less
at highest vacuum
See Sees. 4.4 and 5.0
of EPA Method 3, or
equivalent for Method
3A
Level
No visible leak
observed at 75 mm (3
in.) H20 for 15 seconds
Pitot tube is level
with no visible
rotation from
perpendicular to flow
< 20° average offset
from perpendicular to
flow
See Sec. 4.0
± 10% of 100%
See Sec. 8.2
          0023A - 26
  Revision 1
January 1995

-------
                                 TABLE 3
                REQUIREMENTS FOR ANALYTICAL PREPARATION,
                  SURROGATE  RECOVERIES AND  SAMPLE BLANKS
       Item
Precleaning
filters
Precleaning
sorbent
Filter and
sorbent surrogate
spikes	
Field blank
Method blank
Reagent blanks
Proof Blank
      Description
Soxhlet extraction
Soxhlet extraction
Isotopically-labeled
compounds
Collect one for every
9 sample runs at each
test location
Prepared at analytical
laboratory (laboratory
blank).   One per
analytical batch
One per lot of solvent
used.  Archive for
possible analysis
One per set of
glassware.  Archive
for possible analysis
(collect only if
requested by Agency)
       Control Limit
Detection limits listed in
Sec. 9.2
Detection limits listed in
Sec. 9.2
70 to 130% recovery
< 5 times the Detection
limits
Criteria decided by
laboratory QA officer
Analyze only if requested
by Agency to determine
source of field blank
confirmation
Analyze only if requested
by Agency to determine
source of field blank
contamination
                               0023A  - 27
                                            Revision 1
                                          January 1995

-------
                                   TABLE 4

                     SAMPLE EQUIPMENT CALIBRATION SUMMARY
Equipment
Primary WTM8 or
DGMa
Sample DGM
Sample DGM
Thermometers,
Thermocouples
Nozzle
Pitot tube
AP gauge (if not
an inclined
manometer)
Balance
Barometer
Procedure
Primary
calibration
Full calibration
Post calibration
Calibration
check
ID calibration
Wind tunnel
calibration or
construction
specifications
verification
See Sec. 2.2,
EPA Method 2
Calibration
check
Calibration
check
Frequency
Every 12
months
Every 6
months
After each
test program
Initially
Before every
test program
Before every
test program
Once/test
program
Initially
Initially
Control Limits
± 1% average
Y, < 2% from Yavg
Ypost < 5% from Yfull
± 2°C (3.6°F) at 3
point cal ibration
from reference
thermometer
Repeated
measurements
± 0.1 mm
(0.004 in.)
Specifications
listed in Sec. 4
of EPA Method 2
Within 5% of
reference at three
readings
Observed weight
< 0.5 g from Class
S weight
< 0.1 in. Hg from
primary barometer
aWTM = wet test meter;  DGM = dry gas  meter.
                                  0023A - 28
  Revision 1
January 1995

-------
sac*
    PCDD/PCDF Sampling Train Configuration
                   Figure 1


                  0023A - 29
  Revision  1
January 1995

-------
     Flow
   Direction
   2 Piece
Configuration
                           28/12 Ball Joint
                            Water Jacket  —
                             Condenser
                         28/12 Socket Joint
                           28/12 Ball Joint


                          Glass Wool Plug

                            Sorbent Trap
                    (20-40 gram Sorbent Capacity)
                            40 RC Glass
                           or Teflon® Frit
                         28/12 Socket Joint
   1 Ptoc«
Configuration
       Condenser and Sorbent Module Configurations
                             Figure 2
                            0023A - 30
            Revision  1
         January 1995

-------
FIGURE 3
 LU

                                        O


                                        i
                                        o
                                        (A
                                        I
                                        E

                                        O

                                        a


                                        n
                                        Q
                                        O

                                        i
                                        a
                                        o
                                        a.
0023A - 31
  Revision 1

January 1995

-------
                                           FIGURE 4
   Container 2
(Front Half Rinse)
                             Container 1
                                (Filter)
      I
  Container 4
(Back Half Rinse)
                             Containers*
                            (Sorbent Trap)
 Concentrate to
1-5mLat<37°C
Add Surrogate
Standards


                Add Internal
                 Standards
              Soxhlet Extract
               f or18 Hours
 Concentrate to
1-2mLat<37°C
                Add Internal
                 Standards
              Soxhlet Extract
               for 18 Hours
              Concentrate and
               Bring to 5 mL
               with Hexane
              Concentrate and
               Bring to 5 mL
               with Hexane
r
Add Alternate
Standards

Archive Half in
Freezer for Possible
Repeat Analysis




~l
Perform Sample
Clean-up and
FracttonaDon

T
Add Recovery
Standards
                                                                     Add Alternate
                                                                      Standards
                                                       Archive Half in
                                                     Freezer for Possible
                                                      Repeat Analysis
                             Analyze with
                            06-5 Column
                          If TCOF is Found,
                         Confirm with 06-225
                           Column Analysis
                           Perform Sample
                            Clean-up and
                            Fractiortation
                                                                                  Add Recovery
                                                                                    Standards
                                                                                       T
                             Analyze with
                             06-5 Column
                                                                                       T
                           If TCOF is Found,
                         Confirm with 08-225
                           Column Analysis
                                                                   Surrogate Standards are added to
                                                                   the sorbent trap prior to sampling
                  PCDD/PCOF Analytical Summary Scheme
                                         0023A  - 32
                               Revision  1
                            January  1995

-------
                                 APPENDIX A
                        RECOMMENDED AUDITING  PROCEDURES

     An  audit  is  an  independent  assessment  of data quality.   Both performance
audits and system audits may be performed.

     Performance  Audit  -  A  performance  audit  is  conducted  to  evaluate
     quantitatively the quality of data produced by the sampling, analysis, or
     the  total measurement  system  (sample  collection,  sample recovery,  sample
     analysis, and data processing).

          Audit  Sample - A  performance  audit sample contains  tetra-  through
          octa-isomers  of  PCDD  and  PCDF.    Audit samples  are  not  normally
          required.

          Performance Audit of the Field Test - A field test performance audit
          may  be  conducted by checking the  dry gas meter for  accuracy using
          procedures located in the Quality Assurance Handbook for Air Pollution
          Measurement  Systems  (EPA 600/4-77-027b).    Performance  audits  on
          thermocouple  readings,  AP  gauges, barometric  pressure gauges  and
          others, may  also  be conducted.

          Performance Audit of Data Processing - The data processing procedures
          may  be  audited  by  requiring the  testing  laboratory   to provide an
          example calculation for one  sample run.  This example calculation will
          include all the calculations used to determine the emissions based on
          the  raw field and laboratory data.

     System Audit  -  A system audit is an on-site, qualitative  inspection and
     review of the total measurement  system.

          The  functions of  the auditor are:

          a)  Observe procedures and techniques  of the field team during sample
              collection and sample recovery; and

          b)  Examine  records of  apparatus  calibrations and   other  quality
              control procedures used in sampling and analytical activities

          When on-site,  the auditor  observes the source test  team's  overall
          performance,  including the  following operations:

          a)  Setting the sampling system and leak checking the sample train and
              pi tot tube;

          b)  Collecting the samples  isokinetically;

          c)  Conducting the final  leak checks; and

          d)  Sample documentation  procedures, sample recovery, and preparation
              of the samples for shipment.
                                  0023A - 33                        Revision 1
                                                                  January 1995

-------
                                  METHOD 0031

            SAMPLING METHOD FOR VOLATILE ORGANIC COMPOUNDS (SMVOC)
1.0   SCOPE  AND APPLICATION

      1.1  Method 0031 is used to determine volatile organic compounds  in gaseous
emissions from a wide variety of  stationary sources including hazardous waste
incinerators.  The following compounds may be determined by this method:
Compound                   Boiling Point (°C)           CAS No.'
Acrylonitrileb
Benzene
Bromodi chl oromethane
Carbon disulfide
Carbon tetrachloride
Chl orodi bromomethane
Chloroform
Chloroprene0
Di bromomethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans- 1 , 2-Di chl oroethene
1,2-Dichloropropane
1 , 3-Di chl oropropene
Methylene chloride
Tetrachl oroethene
Toluene
1,1,1-Trichloroethane
1 , 1 , 2-Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
8 Chemical Abstract Services
T r\ r> 1.1 <^ ^ SMA r>*\Tiil^^n*i+w ^»*/J vw
77
80
87
46
77
119-120 (? 748 mm Hg
61
59
97
57
83
32
48
96
106 @ 730 mm Hg
39
121
111
75
113
87
24
Registry Number.
p\*\s*+^lf44'V/ f\ £ 'V It 4 <•• /•• sxrvirt f\t I
107-13-1
71-43-2
75-27-4
75-15-0
56-23-5
124-48-1
67-66-3
126-99-8
74-95-3
75-34-3
107-06-2
75-35-4
156-60-5
78-87-5
542-75-6
75-09-2
127-18-4
108-88-3
71-55-6
79-00-5
79-01-6
75-69-4

   some stationary sources.
c  Reactive compound; may interact with the test matrix.


      1.2   Method  0031  may be used to  prepare  volatile  organic compounds that
have a boiling point between  -15eC and 121°C.   Field application for volatile
organic compounds with boiling points  less  than 0°C should be supported by data
obtained from laboratory  gaseous  dynamic  spiking and gas chromatographic/mass
spectrometric (GC/MS) analysis according to Methods 5041  and 8260 to demonstrate
the efficiency of the sampling and analysis method.
                                    0031-1                          Revision 0
                                                                  January 1995

-------
      1.3   The  method  is not  applicable to  participates  or  aerosols  since
isokinetic  sampling  is  not  performed.    Isokinetic  sampling  is not  required
because the volatile  organic  compounds are  in  the gas  phase  when  they  are
sampled.  Dynamic spiking may require  using reduced sample volumes collected at
flow rates between 250 and 500 mL/min, for a total  sample volume of 20 L.

      1.4   Application  of Method 0031  is  not restricted to those compounds  in the
target analyte list, however,  detection limits have  been determined for these
compounds and acceptable  method performance data have been obtained.  Method 0031
may also be  applied to the compounds  listed in Table  1  if extra care is taken
because of the high volatility of these compounds.

      1.5   Method  0031  is generally not  applicable to polar water-soluble  and
reactive  volatile organic  compounds.   Examples of  polar water-soluble  and
reactive compounds are  shown  in Table  2.  Other examples where Method 0030  (VOST)
sampling  and analytical   methodology  has been  used  inappropriately  include:
bromoform  (boiling  point 137°C,  above  the  maximum  limit  allowed  by  the
methodology),  ethylbenzene   (136°C),   1,2,3-trichloropropane (156"C),  xylenes
(~1408C),  styrene (146eC), 1,1,2,2-tetrachloroethane (146°C at  746  mm Hg),  and
the dichlorobenzenes (~ 175eC).  Although successful  analysis  for these compounds
can be demonstrated by  spiking sorbent tubes, the compounds will not be collected
quantitatively at  the  upper  temperature limit for the  operation of the SMVOC
train.

      1.6   This  method  is applicable   to  the determination  of volatile organic
compounds in the gaseous  effluent of  stationary sources  such as hazardous waste
incinerators with an upper concentration limit per compound  in the emissions of
approximately 1.5 parts  per million (ppm).  Method 0031  is not  appropriate for
gaseous  volatile  organic compound   concentrations  above  this  limit,  since
saturation  of the  analytical  system  or  compound  breakthrough in the field may
occur.   Modifications of analytical  methods to  reduce the concentration of
compounds  entering the  gas  chromatograph/mass  spectrometer (GC/MS),  such as
splitters or dilutions, may prevent saturation of the analytical  system, but the
analytical data are not  accurate if breakthrough has occurred  during sampling.
The analysis of screening samples or  distributive volume samples is recommended
to prevent analytical system saturation  when high analyte concentrations  may be
encountered.

      1.7   The  sensitivity  of  this   method  is  dependent  upon  the  level  of
interferences  in  the  sample matrix  and the  presence of  detectable  levels of
volatile organic  compounds in  the  blanks.  The  target detection limit of this
method is 0.1 ng/m3 (ng/L) of gaseous effluent.  The upper end  of the range of
applicability of this method is limited  by breakthrough  of the volatile organic
compounds on the sorbent  traps used to collect the sample and the ability  of the
analytical  system to  respond within   the  linear  range of the  instrumentation.
Laboratory method development data have  demonstrated a range of 0.1 to  100 jiig/m3
(ng/L)  for  selected  volatile organic compounds collected on a  set  of sorbent
traps  using  a total sample volume of  20 L or less (see Sec. 2.3).

      1.8   The  SMVOC is  designed to be  operated  at  a  sampling  rate  of 1 L/min
with  traps  being replaced every  20  min for  a  total  sampling  time  of  2 hrs.
Analysis of the  traps  is carried out by thermal  desorption purge-and-trap gas
chromatography/mass  spectrometry  (see Methods 5041  and 8260).   Traps  may be


                                    0031-2                          Revision 0
                                                                  January  1995

-------
analyzed  separately  or combined  onto one  trap  to improve  detection limits.
Additional flow rates and  sampling times are acceptable.  For  example, when less
than maximum detection ability is  needed,  it is acceptable to operate the SMVOC
at 0.5  L/min  for a total  of three  40-minute  periods  (two-hour total  sampling
time).  In this example,  a two-hour  sampling time  is maintained, but the number
of sampling tubes  which must be changed in the field  is minimized,  as is the
number of analyses which must be performed.

NOTE:   The  SMVOC sampling  train may be operated no slower than 0.25 L/min, and
        no faster than  1 L/min.

      1.9  This method  is  restricted to  use by,  or under close supervision of,
trained analytical personnel  experienced  in sampling volatile organic compounds
in air.  Each analyst must  demonstrate  the ability to generate  acceptable results
with this method.
2.0   SUMMARY  OF  METHOD

      2.1   This method employs a sampling module and meter box to withdraw a 20-L
sample of effluent gas containing volatile organic compounds from a stationary
source at a flow rate of 1  L/min,  using  a glass-lined probe heated to 130 + 5°C
and a sampling method for volatile organic compounds (SMVOC) train.

      2.2   The gas stream is cooled to  20°C  by passage through  a water-cooled
condenser and volatile organic  compounds are collected on a set of sorbent traps
(Tenax®-GC/Tenax®-GC/Anasorb®-747).   Liquid  condensate  is collected  in  an
impinger placed between  the two Tenax®-GC traps and the Anasorb®-747 trap.  The
first and second traps contain  1.6 g  of  Tenax®-GC each  and the third trap (back
trap) contains 5.0 g  of  Anasorb®-747.   A total number  of sorbent tube sets to
encompass a total sampling  time of 2 hrs is  collected:   i.e.,  if a sampling rate
of 1  L/min  for  20 minutes is  used,  a total  of six sorbent  tube  sets  will  be
collected in 2 hr of sampling.

      2.3   Alternative conditions  for sample collection may be used, collecting
a sample volume of 20  L or less  at a flow rate reduced from 1  L/min.  (Operation
of the SMVOC under these  conditions is referred to as SLO-SMVOC.)  The SLO-SMVOC
may be used to collect 5 L  of sample  (0.25 mL/min for 20 min) or 20 L of sample
(0.5  L/min  for  40 min)   on  each set of  sorbent  tubes.   These  smaller sample
volumes collected  at  lower  flow  rates  should be  considered when  the boiling
points of the volatile organic  compounds of interest are below O'C  (see Table 1)
to prevent breakthrough.  Refer to Sec. 2.2  for the  total  number of tube sets
collected per run.


3.0   INTERFERENCES

      3.1   Interferences  are  encountered in  the analytical methodology and arise
primarily from background contamination of sorbent traps prior to or after sample
collection. Other  interferences may arise from exposure of the sorbent materials
to solvent vapors prior  to assembly and  exposure  to significant  concentrations
of volatile organic compounds  in  the ambient  air at  a  stationary source  site.
To avoid  or  minimize the  low-level  contamination of train components  with
volatile organic  compounds, care should be taken to avoid contact of all interior

                                    0031-3                           Revision 0
                                                                  January 1995

-------
surfaces or train components with  synthetic  organic  materials  such  as organic
solvents, and  lubricating and  sealing  greases.   Train  components  should  be
carefully cleaned and conditioned according to the procedures described in this
protocol.   The  use  of a  sealed/enclosed  sampling  train  is suggested  but not
required (for example, a  purged  glove bag may be used).  The use of blanks (Sec.
6.6) is essential to  assess the  extent  of any  contamination.   Refer to Method
5041 for additional  information on analytical interferences.

     3.2  If the emission source has  a  high  level  of organic  compounds in the
emissions matrix  (for example,  hydrocarbons  present at levels of  hundreds  of
ppm), the presence of these volatile  organic compounds may  interfere with the
performance  of  the   SMVOC  analytical   methodology.    If the  probability  of
saturation of the analytical instrumentation  exists, preliminary SMVOC screening
samples with distributive volumes may be necessary to help ensure that valid and
usable data will  be  obtained.   To perform sampling  according  to  distributive
volumes, samples of different volumes are collected (typically  5 L, 10 L, and 20
L) to verify that analyte  concentrations are  IX,  2X,  and 4X.  The concentrations
of artifacts produced by the sorbent will  not vary with sample volume.


4.0  APPARATUS  AND MATERIALS

     4.1  Sampling train  - A schematic of  the principal components of the SMVOC
is shown in  Figure 1.  The SMVOC  consists of a heated glass-lined probe, followed
by an  isolation valve and charcoal trap,  a  water-cooled glass condenser, two
sorbent tubes containing Tenax®-GC (1.6 +  0.1 g each), an empty knock-out trap
for condensate  removal, a second water-cooled glass condenser,  a  third sorbent
tube containing Anasorb®-747  (5.0 g  ± 0.1  g),  a silica gel  drying  tube,  a
calibrated rotameter, a sampling pump, and a dry gas meter.   The  vacuum during
sampling and for leak-checking is monitored by pressure gauges which are in-line
with and downstream  from the silica gel  drying tube.  The components of the
sampling train  are described below.

          4.1.1  Probe   -  The  probe  is  made of   stainless  steel  with  a
     borosilicate or  quartz glass liner.   The temperature  of  the  probe  is
     maintained at  130'C ± 5'C or higher,   but  not  so high that  the sorbent
     temperature exceeds  20"C.    A  water-cooled  probe  may  be  necessary  at
     elevated  source temperatures to protect the probe  and meet  the required
     sorbent  temperature  maximum.   Isokinetic sample  collection  is not  a
     requirement for  the use of  SMVOC since the compounds  of interest  are in the
     vapor  phase  at the point of sample  collection.   No nozzle is required, but
     a plug of  clean  quartz wool (approximately 2.5 cm. (1 in.)) is inserted in
     the probe  to remove  particulate matter.

NOTE:     No  stainless steel  components should be in contact with  the sample
          stream.

          4.1.2   Isolation valve  - The  isolation valve is a greaseless stopcock
      (0.25  in.  outer diameter stem is recommended) with a  glass bore and sliding
     Teflon® plug with Teflon®  washers  (Ace  Glass 8193 or equivalent).

          4.1.3  Condensers -  The condensers (Ace  Glass 5979-14 or equivalent)
     must be of sufficient capacity to cool the gas stream to 20°C or less prior
     to passage  through  the  first sorbent   tube.   The top connection of the

                                    0031-4                          Revision 0
                                                                  January 1995

-------
condenser must form a  leak-free,  vacuum-tight  seal without  using sealing
greases.  Solverall® tube fittings and screw caps with Solverall® washers
(i in. OD, or equivalent) are recommended.

     4.1.4   Sorbent tubes - See  Figure 2 for  a diagram of  a SMVOC tube.

             4.1.4.1   The first and second tubes of  a three-tube  set  of
     sorbent tubes  should each be packed with 1.6  ± 0.1 g of  Tenax®-GC
     resin and the third  tube of  the  set  should  be packed with  5.0 ±  0.1
     g  of Anasorb®-747.   The  tubes  should  be marked with an  arrow  to
     indicate the direction of flow during sampling.

             4.1.4.2   The sorbent tubes  are  glass tubes  with  approximate
     dimensions of  10  cm x 1.6 cm  ID.   The  tube is  a single glass  tube
     which has the ends  reduced in  size  to accommodate a  i-in.  Swagelok®
     fitting.  The sorbent is held in place by unsilanized clean glass wool
     at each end of the sorbent layer.  Threaded end caps are placed on the
     sorbent tube after  packing with  sorbent to  protect  the sorbent  from
     contamination during storage  and transport.  In order to minimize tube
     breakage, fittings  are finger-tight  plus  an  additional quarter  of a
     turn.  Ceramic-filled Teflon®  ferrules  (Supeltex M2A or  equivalent)
     are used for tubes.  Graphite ferrules (Supeltex M4 or equivalent) are
     used if reconditioning of the tubes is necessary. The  Swagelok® end
     caps should be finger-tightened with the ferrules in place so that the
     entire cap assembly  may be turned as a unit.  In order  to  seal  the
     assembly and avoid glass breakage, the cap assembly  should  be pushed
     to the end of the glass and then backed off slightly before tightening
     the cap with a wrench one quarter of a turn.   Backing the cap assembly
     off  from  the end  of the  tube will  prevent chipping,  cracking,  or
     breaking of the glass.

             4.1.4.3   The sorbent  tubes  are  placed  in  transport  tubes
     (capped culture tubes with glass wool and  charcoal)  for shipment.   A
     layer of clean charcoal  is  placed  in  the bottom of the transport  tube
     to absorb any volatile organics in the air in the transport tube.   A
     plug of cleaned glass wool (approximately  2.5 cm. (1 in.))  is placed
     above the charcoal.  The SMVOC  tube, with  both ends capped,  is placed
     in the transport tube, and a plug of cleaned glass wool  (approximately
     2.5 cm. (1 in.))  is  placed on  top of the  SMVOC  tube.   The  two glass
     wool plugs cushion the SMVOC tube during  shipping. The transport tube
     is then sealed tightly with a Teflon®-!ined  screw cap.

     4.1.5   Metering  system -  The  metering  system for SMVOC  consists  of
a vacuum gauge, a pump,  a calibrated rotameter  for monitoring the sampling
flow rate, a dry gas meter (2% accuracy, with a minimum resolution of 0.01
L) at the required sampling rate, needle valves, and  a temperature readout
device.  Provisions should be made  for monitoring the temperature  of the
sample gas stream between the first condenser and the first  sorbent tube,
since  this  temperature should  not exceed 20°C.   The temperature  can  be
monitored by placing a  thermocouple on the exterior glass surface  of  the
outlet from the first condenser.  The temperature at  that point  should  be
less than 20°C.  If the cooling  is  not sufficient,  an  alternative condenser
providing the necessary cooling capacity must  be  used.
                               0031-5                          Revision  0
                                                             January  1995

-------
           4.1.6    Sample  transfer  lines  - All sample transfer lines connecting
      the  probe to  the SMVOC shall  be less than 1.52 m. (5 ft.) in length.  The
      sample  transfer lines shall be heat-traced Teflon® or glass maintained at
      130  ± 5°C.    Connecting  fittings must  be  capable of  forming  leak-free,
      vacuum-tight  connections without the use  of sealing grease.   All  other
      sample  transfer lines used with the  SMVOC shall be Teflon® with connecting
      fittings that are  capable  of  forming leak-free,  vacuum-tight connections
      without the use of sealing grease.   These sample transfer lines should not
      be reused at  other emission sources.

      4.2   Solverall® washers  -  All washers or gaskets used  in  SMVOC shall be
Teflon®-coated (Solverall® washers or equivalent; } in.  stainless steel Swagelok®
fittings with Supeltex  M2A  ferrules may  also be  used).   Prior  to  use,  these
gaskets should  be  ultrasonically-cleaned with  methanol  and  air-dried  in  a
contained/isolated organic vapor-free area.   Gaskets should be stored in clean,
screw-capped containers prior to use.

      4.3   Glass wool - Glass  wool  shall  be Soxhlet-extracted for 8 to 16 hours
using methanol,  and oven  dried  at  110°C  before  use.   Glass  wool  should not be
silanized  to prevent contamination  of samples with  siloxanes.   Quartz wool is
recommended for high temperature applications.

      4.4   Cold packs/ice  -  Ice  or  any commercially-available reusable liquids
or gels that  can be frozen  repeatedly are  acceptable. These reusable liquids are
typically  sold in  plastic  containers as "Blue  Ice" or "Ice-Packs".  Enough cold
packs or ice should be used to maintain tubes less than 10°C.   If  ice is used as
a coolant  for the tubes, the  tubes should be  shielded  from direct contact with
the ice so they will not become wet when  the  ice melts.  Use of dry ice (solid
C02)  for cooling  tubes  should be  avoided;   the sorbent  tubes take  up carbon
dioxide as the solid coolant vaporizes and the analytical  system  is vented when
the tubes  are desorbed and analyzed.  The  tubes should  not  be  stored at freezing
temperatures, since  the  seal  between the glass  and Teflon®  fittings  will be
compromised  and diffusion of  volatile organic compounds  into the  sorbent may
occur.

      4.5   VOA  vials -  40-mL  glass  vials with  Teflon®-!ined  screw  caps  are
required for recovery of condensate.

      4.6   Teflon®  squeeze bottles  - Teflon®  squeeze bottles  should  be washed
with a solution  of a laboratory  detergent, rinsed with  hot tap water, then with
distilled  water,  then rinsed  with clean purged water prior to use.


5.0   REAGENTS

      5.1   Reagent  grade chemicals  shall  be used  in  all  tests.  Unless otherwise
indicated, it is  intended  that all  reagents shall conform  to the specifications
of the Committee  on Analytical Reagents of the American Chemical  Society, where
such  specifications are available.   Other grades may  be  used,  provided  it is
first ascertained that  the reagent  is of  sufficiently high purity to permit its
use without lessening the  accuracy  of the  determination.  It is recommended that
blanks be  taken of all  reagents used in testing.
                                    0031-6                          Revision 0
                                                                  January 1995

-------
      5.2   2,6-Diphenyl-p-phenylene  oxide polymer  (Tenax®-GC,  35/60  mesh,  or
equivalent).

           5.2.1    New Tenax®-GC is Soxhlet-extracted for 24 hours with methanol.
      The  Tenax®-GC is  dried for 6 hours in a vacuum  oven  at 50°C before use.
      Thermal conditioning  (Sec. 7.1.1) of the  Tenax®-GC should  be  done prior to
      blanking.

           5.2.2    If reuse  of Tenax®-GC  is  necessary,  the  polymer may  be
      extracted sequentially with methanol and pentane, dried in a vacuum oven,
      and thermally reconditioned as described above.  However, reused tubes must
      meet  the same criteria for cleanliness  as new  tubes.  Reuse of sorbents is
      not  recommended.   Common practice  in laboratories where  SMVOC  tubes are
      prepared  commercially  or where  SMVOC  sampling and  analysis   are  done
      extensively  is  not  to reuse  sorbents.

      5.3   Anasorb®-747  - New Anasorb®-747 is  used as  it  is  received  from the
manufacturer without preparation  other  than  thermal  conditioning pending  a
Quality Control  check  (Sec.  7.1.1).   Anasorb®-747 must  not be  reused.   The
Anasorb®-747 should  not  be extracted with organic  solvent  prior to  use  as  a
sorbent in the SMVOC.

      5.4   Silica  gel -  Indicating type,  6-16 mesh.   New silica gel may be used
as received from the vendor.  Silica gel  should  not be reused for SMVOC.

      5.5   Methanol,  CH3OH  - The methanol  used  for extracting the Tenax®-GC and
glass wool should be pesticide grade or equivalent.

      5.6   Organic-free  reagent  water - All  references to  water in this method
refer to organic-free reagent water,  as defined  in  Chapter One.


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   SMVOC glassware  cleaning - All glassware should be cleaned using the
following steps.

           6.1.1    Sonicate in  an ultrasonic bath for  1 hour  in a solution of a
      laboratory detergent  such  as Alconox®.

           6.1.2    Rinse  with copious amounts of hot  tap  water to  remove all
      detergent residue.

           6.1.3    Rinse  three  times  with HPLC grade water.

           6.1.4    Oven  dry at  110°C.

           6.1.5    Cap  for  shipment using Teflon® tape or aluminum foil.

      6.2   Assembly

           The assembly  and packing of  the sorbent  tubes should be carried out
      in an area free of volatile organic material,  such as a laboratory in which


                                    0031-7                          Revision  0
                                                                  January 1995

-------
     no organic  solvents are handled or stored and in which the laboratory air
     is  charcoal filtered.  Alternatively,  an  air-tight sealed glove  box is
     suggested.

     6.3  Tenax®-GC  tubes

          6.3.1    The Tenax®-GC glass tubes and metal tube parts  are cleaned and
     stored  (see Sec.  6.1).   Ferrules are  discarded but  the  metal  unions are
     cleaned by  sonication in methanol.   Tenax®-GC (1.6 ±  0.1 g) is weighed and
     packed  into each  of  the  first  two  sorbent tubes which  have  unsilanized
     cleaned glass wool in the downstream end.  The Tenax®-GC is held in place
     by inserting unsilanized cleaned glass wool.  Each tube should be marked,
     using an  engraving tool,  permanent  marker  or diamond-tipped  pencil,  with
     an arrow  to indicate  the  direction  of sample flow during sampling,  and a
     serial number.

          6.3.2    Conditioned sorbent tubes are capped  and placed on cold packs
     or  ice  for storage and transport.   The temperature of  the  tubes  during
     storage and transport is  maintained at a temperature  of  less  than  10°C.
     Conditioned tubes should be held for no more  than 14 days before sampling,
     to prevent  the  possibility of contamination.

     6.4  Anasorb®-747 tubes - Anasorb®-747 (5.0 ± 0.1 g)  is weighed and packed
into the third  sorbent tube which also has unsilanized cleaned glass wool in the
downstream end.   The Anasorb®-747 is held in place by  inserting  unsilanized
cleaned glass  wool.   Special care should  be taken to conspicuously  mark the
Anasorb®-747 tube  with an arrow  to indicate  the direction  of  flow  during
sampling,  and a  serial  number.

     6.5  Sample collection

          6.5.1    After leak checking (see Sec. 6.5.3) but before the initiation
     of sample collection,  the  probe  shall be purged with  stack gas.  This purge
     can be accomplished by attaching a pump to the isolation valve upstream of
     the  first  condenser  and  drawing  stack gas  through  the  probe via  the
     isolation  valve,  so  that the  probe  is  purged  of ambient  air at  the
     initiation  of sample  collection.

          6.5.2    Sample collection  is accomplished by  opening the valve at the
     inlet  to  the first condenser  (see  Figure  1),  turning on the  pump,  and
     sampling  at a rate of  1  L/min (or slower rate,  if desired, according to the
     guidelines  for SLO-SMVOC)  for 20  minutes  (or  an   appropriately  longer
     period, if  slower sampling rates are used).  The volume of sample for any
     set  of  traps  should  not exceed 20  liters.   The end caps  of  the sorbent
     tubes  should  be placed  in  a clean  screw-capped  glass  container  during
     sample collection to  prevent contamination.

          6.5.3    Following  completion of sample collection, the SMVOC is leak-
     checked a second time  at the highest  vacuum encountered  during the sampling
     run  to minimize  the  chance  of vacuum desorption  of  volatile  organic
     compounds  from  Tenax®-GC.   The sample  is considered invalid  if the  leak
     test does  not meet specifications.   The train is  returned  to atmospheric
     pressure  and the  set of sorbent  tubes is  removed.  The end  caps  are
     replaced  and the  tubes are  placed in  an  organic-free environment  and

                                    0031-8                          Revision 0
                                                                  January 1995

-------
     maintained at a temperature less than 10°C  for  storage and transport.  The
     set of  tubes  and  any  condensate collected  (see Sec.  6.5.4)  are placed in
     self-sealing  plastic  storage bags.

           In the laboratory, tubes are maintained in  a clean, organic vapor-free
     environment at  a  temperature less  than  10°C  until  analysis.   The maximum
     storage time between sampling and analysis  of the tubes should be 14 days.
     The rate  of  loss  of sorbed volatile organic compounds from  the tubes is
     both compound-specific and source-specific.  A  14-day period  is chosen for
     the  holding  time  before  analysis to  provide  a reasonably  conservative
     guideline for quantitative analysis of the volatile  organic compounds which
     have  been sampled.

NOTE:        To prevent breakage and/or  loosening of the seals  at the end of the
             tubes,  SMVOC  tubes  should  not be stored in a  freezer or over dry
             ice.    A  solvent-free  refrigerator  (no  cooler  than  4°C)  is
             appropriate for storage of the tubes until analysis.

           6.5.4    The   condensate  is  recovered  by transferring  any  liquid
     contained  in  the  knock-out trap  to a 40-mL  VOA vial  and  rinsing  the
     knock-out trap  three  times  with a  minimum  volume  of  organic-free reagent
     water  (Sec.  5.6)  and adding the rinses to the VOA vial.   If necessary,
     water  should  be added to eliminate  headspace  in  the vial.    If  there is
     sufficient condensate to fill more  than one  vial, two vials should be used.
     The VOA vials containing the condensate  are placed, with the set of tubes,
     in a self-sealing plastic storage bag and maintained at a temperature less
     than  10°C  for storage and  transport until  analysis.   The  condensate is
     analyzed by Method 8260.   Refer to Method  8260 for details  on analytical
     procedures.

          6.5.5    A  new set of  tubes  is placed  in the  SMVOC,  the  SMVOC is
     leak-checked,  and the  sample  collection  process  repeated   as  described
     above.  Sample collection continues until  sufficient samples to encompass
     a two-hour sampling period  have  been collected.   If  samples  are taken at
     a sampling rate of 1  L/minute,  a two hour  sampling period will  result in
     the collection  of six sets of  tubes.  If  SLO-SMVOC procedures  are  used,
     fewer than six  sets of tubes will  be sampled over  a two-hour period.

     6.6  Blanks

          6.6.1    Field blanks - Blank Tenax®-GC  and Anasorb®-747  tubes  are
     attached to the sampling train  while  the train  is leak-checked.  The tubes
     are removed and stored with the sample tubes.   At least one field  blank
     should be collected for every two-hour sampling period.

          6.6.2   Trip  blanks  - At least one set of  blank tubes (two Tenax®-GC,
     one Anasorb®-747)  should  be included  with each shipment of tubes to  a
     stationary source sampling site. These trip blanks should be treated like
     any other tubes  except that the end caps will  not be removed during storage
     at the site.   This set of  tubes  should be analyzed to  assess  contamination
     which may occur during storage  and shipment.

          6.6.3    Laboratory blanks  - One  set of  blank tubes (two Tenax®-GC,  one
     Anasorb®-747) should remain in  the laboratory using the method of storage

                                    0031-9                          Revision  0
                                                                  January  1995

-------
     which  is used  for  field  samples.   ihese  laboratory  blanks  should  be  from
     the  same batch of  sorbent as used for the field  blanks, trip  blanks  and
     collected   samples.     If  the  field  and  trip   blanks   contain   high
     concentrations  of  contaminants  (e.g.,  greater  than 2 ng of  a  particular
     volatile organic  compound],  the  laboratory  blank should be analyzed  in
     order  to identify  the source of contamination.


7.0  PROCEDURE

     7.1  Tube condition!ng

          7.1.1    In  a  desorption oven,  the sorbent tubes are  connected  to a
     source of organic-free nitrogen.  Nitrogen is passed through each  tube at
     a flow rate of 80-100 mL/min while the tubes  are heated.  Anasorb®-747 is
     thermally conditioned for 18-24 hours  at 300°C, under a nitrogen flow rate
     of 80-100 mL/min.   Tenax®-GC is thermally conditioned at  220°C for  8-12
     hours  at a nitrogen flow rate of 80-100 mL/min.  The actual  length  of time
     required  for  the  conditioning period  may  be  determined   based  on  the
     adequacy of  the  resulting  blank checks of the  conditioned  tubes.   Method
     5041 (modified to use a sorbent desorption temperature of 250°C) and Method
     8260 may be used to perform a blank check  of  each set of sampling tubes to
     ensure cleanliness.

          7.1.2    An acceptable blank level  is  less  than  or equal to (<} Method
     Detection Limits for Method 5041/8260 (see Method 8260 for Method Detection
     Limits).  A general guideline  of analyte values  less  than 2  ng  for  any
     volatile organic compound may be used as  a criterion of cleanliness.

          7.1.3    After conditioning, tubes are sealed  and placed on cold packs
     or  ice (maintained  at  a temperature  less than  10°C)  until sampling  is
     completed.   Conditioned tubes  should  be  held  for  no  more than  14  days
     before sampling, to  prevent  the possibility  of contamination.

     7.2  Pretest  preparation

          7.2.1    All  train  components  should be  cleaned  and assembled  as
     previously described.  A dry  gas meter should be calibrated  within  30 days
     prior  to  use,  using  a  standard  orifice,  or other approved  calibration
     device/meter.

          7.2.2    The SMVOC  is assembled according to the schematic  diagram in
     Figure 1.   Cooling water should be circulated to the condensers  and  the
     temperature  of the  cooling  water  must   be  iow  enough  to maintain  the
     temperature  of  the  gas entering the sorbent  below 20°C.

     7.3  Leak-checking

          7.3.1    To leak-check the entire  train,  it is necessary to leak-check
     from  the  probe  to the   pump.   In order  to  adequately represent  actual
     sampling conditions, a leak-check  should  be performed with the pump on and
     the  leak rate measured  in liters  per  minute  (Lpm)  on the  dry  gas meter.
     After  the desired  vacuum is  reached, the  pump  is  isolated  from the  train
     to check for  leaks.

                                   0031-10                          Revision 0
                                                                  January 1995

-------
           7.3.2   Ensure that all connections are tight and that the train is
      assembled correctly with sorbent  cartridges  properly assembled  and  in the
      right direction for  sampling.   Seal the  end  of the  probe  and turn the
      isolation valve to  the  sample/open  position.  Turn on  the pump  and  adjust
      the  vacuum to  25.4  cm above  normal  operating pressure  (38 cm Hg should be
      sufficient as  12.7  cm or less  is  normal).  Prior to leak-checking,  verify
      that the fine  adjust valve on the meter box is partially opened and that
      the  coarse adjust valve  is  almost  closed  to prevent  backflushing of any
      condensate during final  leak checks as the valves will  need to be adjusted
      to increase vacuum rather than decrease  vacuum.  Allow the  rotameter  on the
      meter box  to  drop  to  zero  and the  pressure  on the  water  column  gauge
      (represents the pressure inside the dry gas meter)  to stabilize.  The pump
      is   isolated by shutting  off  the  needle  valve.   Record the  leak rate
      directly from  the dry gas meter and time for  one  minute using a  stopwatch.
      The  leak rate  must be less than 0.02 Lpm.   This value should be  sufficient
      as it is less  than  1% of the sample rate.

           7.3.3  Upon completion  of  the leak check,  turn off the pump and
      release  the pressure/vacuum  in the train by turning the isolation valve to
      the  purge position  and  allowing  ambient air  (filtered with  charcoal  or
      equivalent) to enter the train.  The initial  leak-check  should be above
      normal operating pressure.   The final leak-check (following collection of
      20 L of sample) should be at  least at the highest  vacuum encountered during
      the  run.

NOTE:         The volume  of air pulled through  the  SMVOC  during leak-checking
              procedures prior to sampling should be  less than 2.5% of the total
              volume sampled.   If  a  volume  greater  than  2.5%  of  the  total
              sampling  volume  is   pulled  through  the  SMVOC  in  obtaining  a
              successful  leak check, the  sorbent  tubes  used during  this leak
              check  must be discarded and a successful  leak check with a minimum
              volume of gas pulled  through the train  must be  obtained with a new
              set of sampling  tubes  in  place.

      7.4   Sample collection - Sample collection procedures are described in Sec.
6.5.

      7.5   Analytical  procedure - Samples are analyzed  by Methods 5041 and 8260.
In these methods, adapted for a three-tube SMVOC,  the sorbent tubes are spiked
with  surrogates, internal  standards are spiked into  the purge  water,  and the
tube(s) thermally desorbed at 250°C under a purge of organic-free helium.  The
tubes may  all  be analyzed individually, or the Tenax® tubes may be analyzed as
a pair with the Anasorb® tube  analyzed separately.  The gaseous effluent from the
tubes is bubbled through purged organic-free reagent water (Sec.  5.6) and trapped
on an analytical sorbent trap in a purge-and-trap unit.   After  desorption, the
analytical sorbent  trap is heated  rapidly and the gas flow from the analytical
trap is directed to  the head of a wide-bore capillary column  (Method 5041) under
subambient  conditions.    The  volatile  organic  compounds   desorbed  from  the
analytical trap are  separated by temperature-programmed  gas  chromatography and
detected by continuously-scanning low resolution mass spectrometry (Method 8260).
Concentrations of volatile organic  compounds  are  calculated from  a multipoint
calibration curve,  using the  method of response factors.   Refer to Method 8260
for details.
                                   0031-11                          Revision 0
                                                                  January 1995

-------
     7.6   Calculations

           7.6.1    The following  nomenclature  is  used in  the calculation  of
sample volume:

     Pbar         =  Barometric pressure at the exit orifice of the dry gas meter,
                   mm (in.)  Hg.

     Pstd         =  Standard  absolute  pressure,  760 mm Hg.


     Tm         =  Dry gas meter  average  absolute temperature, °K (°R)


     Tstd         =  Standard  absolute  temperature,  293°K (528°R)


     Vm         =  Dry gas volume measured  by dry gas meter,  dcm (dcf)


      V
       mi.idi
                 =  Dry  gas  volume measured  by  dry  gas  meter,  corrected  to

                   standard  conditions, dscm (dscf)

     y           =  Dry gas meter  calibration factor

           7.6.2    The volume of gas sampled is calculated as  follows:


                                   T  P           V  P
                       u   _ ii      std bar _  i/     m  bar
                        m.,rf   nn '  T  r\      1 '   T
                                    m  std
     where:

                   KT  =  0.3858°K/mm Hg for metric units or

                   K,  =  17.64°R/in.  Hg for English units.


           7.6.3    The concentration of volatile  organic  compound  (CPD)  in the
     stack sample (Cg)  is  calculated  as  follows:


       Total weight of CPD in sample (i.e., analytical  measurement from VOST tubes)
  \j    — — --    ~~"   ••— - -             .,— .
   9                   Volume of sample at  standard conditions, dscm
                                    0031-12                          Revision 0
                                                                   January 1995

-------
8.0  QUALITY CONTROL

     8.1   Prior  to actual  sampling  on-site,  all  of the  applicable sampling
equipment should  be thoroughly checked to ensure that each component is clean and
operable.  Each of the equipment calibration data forms should be reviewed for
completeness and adequacy to ensure  the  acceptability  of the equipment.   Each
component of the  sampling system should  be  carefully  packed for shipment.  Upon
arrival on-site,  the equipment should  be unloaded, inspected for possible damage,
and then assembled for use.

     8.2   The  following  quality  control  (QC)  checks  are  applicable to  the
sampling procedures:

           8.2.1    Each  sampling train  must  be visually  inspected  for proper
     assembly  before  every  use.

           8,2.2    All  sampling  data  should be recorded on standard data forms
     which may serve  as  a pretest checklist.

           8.2.3    The temperature measurement system should be visually checked
     for damage  and operability by measuring  the ambient  temperature.

           8.2.4    All  sampling data  and  calculations  should be  recorded  on
     Preformatted  data  sheets.

           8.2.5    All  glassware for  SMVOC should be cleaned  according to the
     procedure in  Sec.  6.1.

           8.2.6    Ten percent  of  the SMVOC tubes should be subjected to GC/MS
     QC measurements.   No analytes should  be detected  at concentrations above
     method detection  limits  in  unused  SMVOC tubes.   If these quality control
     tests are performed by the manufacturer, documentation should be obtained
     from  the  commercial supplier and retained.

           8.2.7    All  cleaned  glassware,  hardware,  and prepared  sorbent traps
     should be kept closed with ground-glass caps or Teflon® tape until assembly
     of the sampling  train  in the field.  The sorbent traps should be recapped
     immediately after  each set of samples is  collected.

           8.2.8    Prior to sampling,  the Tenax®-GC and Anasorb®-747 tubes should
     be  spiked with  the compounds  of  interest to  ensure  that  they can  be
     thermally desorbed under laboratory conditions.  This spiking is necessary
     but not sufficient.  The compound must  still be sampled from the  source.

           8.2.9    Assembly  and recovery  of  the sampling trains must be performed
     in  an environment as  free from uncontrolled  dust and  solvent  vapors  as
     possible.

           8.2.10   Blanks  (field, trip,  laboratory) must  be collected.

           8.2.11   The  entire  sampling train  should  be  leak-checked before and
     after each run.   If the sampling train is moved from one sampling port to
     another during a  run, the train should be leak-checked before  and after the
     move.

                                   0031-13                          Revision 0
                                                                  January 1995

-------
     8.2.12  Dry gas meter readings, temperature readings, and pump vacuum
readings  should  be  made  during sampling  and recorded  in  intervals  no
greater than 5 minutes.

     8.2.13  Sorbent traps should be used for  sampling within two weeks of
preparation.

     8.2.14  During  sample collection, the gas stream temperature  at  the
inlet to the first sorbent trap must be maintained at or below 20°C.

     8.2.15  All sample traps should be stored under refrigeration  or on
ice or cold packs (temperature maintained  less than  10°C)  until  ready  for
analysis.

8.3  QC for analytical procedures

     8.3.1   Calibration standards should be prepared at five concentration
levels for  each  analyte of  interest.   Compounds  of interest,  surrogate
compounds,  and  internal  standards are  spiked into the  purge water  for
generation of a multipoint calibration curve.   When  samples  are  analyzed,
surrogate  compounds  are  spiked  onto  the  sampling  tubes  using  flash
vaporization techniques (Method 5041), but internal standards are  spiked
into the purge water.  Response factors for each  compound  are calculated
and  these  response  factors  are  used  for the calculation of analytical
results.    Refer  to Methods  5041  and  8260  for  detailed  analytical   QC
procedures for analysis of samples.

     8.3.2   To  establish  the  precision  and  accuracy  of the  analysis,
triplicate  paired  Tenax®-GC tubes  should  be  spiked  with analytical
surrogate volatile organic compounds  using  flash evaporation  and analyzed
immediately following the initial  calibration  and  before  sample  analysis.
The  spiking  level  should be  at the expected  level  of volatile  organic
compounds in the stationary  source.  The spiking standard must be prepared
from stock standards  separate from those used for calibration.   Recovery
for each volatile organic compound and surrogate  should be within  50% to
150% of spiked value.  The relative standard deviation associated with each
analyte should be less than 25 percent.

     8.3.3   The average recovery from the initial precision and accuracy
determinations  should be  used  as  an  acceptance  criterion  for  sample
results.    The  surrogate recovery in each  sample  should be within  three
standard  deviations  of the  average  recovery obtained  from the  initial
precision and accuracy determinations.

     8.3.4   An  EPA  performance audit  should  be completed during  a  trial
burn as a check on the entire SMVOC system.  The audit results should agree
within 50%  to  150% of the expected value  for each specific compound  of
interest.  This  audit  consists of collecting  a gas  sample  containing  one
or more  volatile organic compounds in the  SMVOC from  an EPA  audit  gas
cylinder.   Collection  of  the audit sample in  the SMVOC may  be  conducted
either in the laboratory or  at the field test site.  Analysis of  the SMVOC
audit sample must  be by  the  same  person,  at the same time,  and with  the
same analytical  procedure as  used for  the regular SMVOC  samples  from  the
field test.

                              0031-14                         Revision 0
                                                             January 1995

-------
9.0  METHOD PERFORMANCE

     See Method 8260.


10.0 REFERENCES

1.   "Field  Test  of a  Generic  Method  for  Halogenated  Hydrocarbons"  (U.S.
     Environmental Protection Agency).  EPA 600/R93/101, June  1993.
                                   0031-15                          Revision  0
                                                                  January  1995

-------
                                    TABLE  1
                      HIGH VOLATILITY ORGANIC COMPOUNDS3
Compound                     Boiling Point
Bromomethane                       4°C

Chloroethane                      12°C

Vinyl bromide                     16°C, at 750 mm

Vinyl chloride                   -13.4°C
aUse  of SLO-SMVOC may be helpful
                                    0031-16                         Revision 0
                                                                  January  1995

-------
                                   TABLE 2
               COMPOUNDS FOR WHICH METHOD 0031  IS  NOT  APPLICABLE
Compounds
Boiling Point
Allyl chloride




Acetone

Methyl ethyl  ketone

Chloromethane
Epichlorohydrin
     45°C
     56°C

     80°C

    -24°C
    116°C
Chloromethyl  methyl  ether     56°C


bis(Chloromethyl)  ether      106°C


Acetonitrile                  82°C

Acetaldehyde                  21°C

Acrolein                     53°C

Methanol                     65°C

Ethanol                       78°C

Isopropyl  alcohol             82°C
Comment
Reactive  compounds;  interacts
with test matrix to yield  poor
recoveries    and   poor
reproducibility

Polar,  water  soluble

Polar,  water  soluble

Reactive  compounds;  interacts
with test matrix to yield  poor
recoveries    and   poor
reproducibility

Not    amenable    to   SMVOC
analytical procedure

Not    amenable    to   SMVOC
analytical procedure

Not    amenable    to   SMVOC
analytical procedure

Polar,  water  soluble

Polar,  water  soluble,  reactive

Polar,  water  soluble,  reactive

Polar,  water  soluble

Polar,  water  soluble

Polar,  water  soluble
                                   0031-17
                                           Revision  0
                                         January 1995

-------
                                               FIGURE 1
           SCHEMATIC OF  SAMPLING METHOD  FOR VOLATILE  ORGANIC  COMPOUNDS (SMVOC) TRAIN
Glass
Wool
Filter
 Stack
           Healed Glass
           Lined Probe
                              Three Way      1/8'OD
                             GlasaATeflon    Teflon Line
                                Valve     A
                                          Condensers
                                    Condensale
                                       Trap
                                                                                 Condenser
                                                                                  Ice Balh
                                                                                        Temperature
                                                                                         Indicators
        f       f
                                                                                                       Exhaust
Meter Box
                                   Sampling Module
                                                 0031-18
               Revision  0
             January 1995

-------
 FIGURE  2
SMVOC TUBE
                      Glass
                      Wool
             0.6 cm 00
 0031-19
  Revision 0
January 1995

-------
                                  METHOD 0040

             SAMPLING OF PRINCIPAL ORGANIC HAZARDOUS CONSTITUENTS
                  FROM COMBUSTION SOURCES USING TEDLAR® BAGS
1.0   SCOPE AND APPLICATION

      1.1  This  method  establishes  standardized  test  conditions  and  sample
handling procedures for the collection of volatile organic compounds collected
from  effluent  gas samples  from stationary  sources,  such as  hazardous  waste
incinerators  and  other  combustion  sources,  using time-integrated  evacuated
Tedlar® bags.  The compounds listed below can be collected by  this method.  This
is a sample collection method and does not directly address the analysis of these
samples.   Gas  chromatography/mass spectrometry (GC/MS)  (Method 8260)  is the
recommended  analytical  technique because of  its  ability to  provide positive
identification of compounds in complex mixtures such as stack gas.
            Compound                                CAS Registry No.
            Dichlorodifluoromethane                              75-71-8
            Vinyl  chloride                                       75-01-4
            1,3-Butadiene                                       106-99-0
            l,2-Dichloro-l,l,2,2-tetrafluoroethane                76-14-2
            Methyl  bromide                                       74-83-9
            Trichlorofluoromethane                              353-54-8
            1,1-Dichloroethene                                   75-35-4
            Methylene chloride                                   75-09-2
            1,1,2-Trichloro-trifluoroethane                       76-13-1
            Chloroform                                           67-66-3
            1,1,1-Trichloroethane                                71-55-6
            Carbon  tetrachloride                                 56-23-5
            Benzene                                              71-43-2
            Trichloroethene                                      79-01-6
            1,2-Dichloropropane                                  78-87-5
            Toluene                                             108-88-3
            Tetrachloroethene                                   127-18-4
      1.2  This method  is  not  applicable  to  the collection of samples in areas
where there  is an explosion hazard.  Substitution of intrinsically safe equipment
or procedures for the equipment or  procedures described in this method will not
be sufficient to  adapt this method  for  use in areas where there is an explosion
hazard.  Additional modifications to the sampling and analytical  protocols may
be required.

      1.3  This method does not employ  isokinetic sampling and therefore is not
applicable to the collection of highly  water soluble volatile organic compounds
contained  in an  aerosol   of  water.   This  method  uses  either  constant  or
proportional rate sampling, depending upon the  extent of the variability of the
emission flow rate.


                                   0040 - 1                          Revision 0
                                                                  January 1995

-------
      1.4   This  method  is  restricted  to  use  by,  or under the close supervision
of, trained analytical personnel experienced  in  sampling  organic  compounds in
air.  Each analyst must demonstrate the ability to generate acceptable results
with this method.

      1.5   Each  compound for which this method can be considered shall meet the
criteria listed  in  Sees.  1.5.1  - 1.5.3, below.   Method  8260  provides boiling
points,  condensation  points  (calculated  from  vapor pressure)  at  20°C,  and
estimated  instrument  detection  limits.   This  method  is  not  limited  to  the
compounds in the target analyte list, however, stability and recovery shall be
demonstrated when  compounds other than those  listed  in Sec.  1.1 are  to be
sampled.

           1.5.1    The  compound  must  have  a  boiling  point  <  121°C.

           1.5.2    The  compound must have a concentration in  the stack gas below
      the condensation  point.

           1.5.3    During  validation  studies,  the loss of the  compound  from a
      Tedlar® bag must  be  less than 20%  over a  72-hour storage time.


2.0   SUMMARY OF  METHOD

      2.1   A representative sample is  drawn from a source  through a heated sample
probe and filter.

      2.2   The  sample  then  passes through  a  heated  3-way valve  and  into  a
condenser where the moisture and condensable components are removed  from the gas
stream and collected in a trap.

      2.3   The  sample  is  collected in  a Tedlar®  bag  held in a  rigid,  opaque
container.

      2.4   The   dry  gas  sample  and   the  corresponding   condensate  are  then
transported together to a GC/MS.   A  mass spectrometer  is most  suited for the
analysis and quantitation  of complex mixtures of volatile organic compounds.  The
total amount of the  analyte in the sample is determined by summing the individual
amounts in the bag  and the condensate.  A flow chart of the procedure is given
at the end of this  method.
3.0   INTERFERENCES

      3.1   The materials from which the Tedlar® bag is constructed may contribute
background hydrocarbon contamination.   Purging the bag with air or  N2 may reduce
the concentration of these hydrocarbons.  Exposure of the bag to direct sunlight
may increase the concentration of these hydrocarbons.  Therefore,  the bag must
be protected from exposure to  sunlight by using an opaque container to house the
bag during sampling and shipping.
                                   0040 - 2                         Revision 0
                                                                  January 1995

-------
     3.2  Components of the source emissions other than the target compounds may
interfere.  Interferants may be  differentiated from the target compounds during
mass spectrometric analysis.

     3.3  Common  problems that can  invalidate  Tedlar® bag sampling  data  and
techniques to remedy these problems are listed in Table 1.

     3.4  Available stability data suggest that this method may not perform well
in sampling streams containing  polar  and reactive  compounds  like  methyl  ethyl
ketone, formaldehyde,  methanol,  1-butene,  and acetone.  The  use  of this method
to sample these compounds needs to be evaluated before sampling.


4.0  APPARATUS AND MATERIALS

     4.1  Tedlar®  bag  sampling  train - A detailed  schematic of the principal
components of the sampling train is shown  in Figure 2.

          4.1.1    The sampling train (Figure 2) consists of a glass-lined probe,
     a heated glass or Teflon®  filter holder and quartz filter attached to one
     of two inlets of a glass and Teflon® 3-way isolation valve  (Figures 3 and
     4).  The second valve  inlet  is used to release system pressure after leak
     checks.  This valve is  connected to a  charcoal  trap, which filters incoming
     air.   The  outlet  of the isolation valve is connected to a glass,  water-
     cooled,  coil-type  condenser  and a glass condensate trap for removal  and
     collection of condensable liquids present in the  gas stream.  A 1/4-in. OD
     x 1/8-in. ID Teflon® transfer line connects the condensate trap  to a second
     3-way  isolation valve  and  the  isolation valve to a Tedlar® bag contained
     in a rigid, air-tight  container for sampling,  storage,  and  shipping.  The
     bag  container is connected to a control console with 1/4-in. OD x 1/8-in.
     ID  vacuum  line by  means  of  1/4-in.  Teflon®  connectors  at each end.   A
     charcoal trap is placed  in the  vacuum  line between the  bag container and
     the  control  console to  protect  the  console and  sampling  personnel  from
     hazardous emissions  in case  of bag rupture during  sampling.

          4.1.2    The vacuum  required to  operate this system is provided by a
     leak-free diaphragm pump contained in the control  console (Figure  5).  When
     the  pump is  turned on, the space between  the   inner  walls of the  bag
     container  and the  Tedlar®  bag  is evacuated, placing  the  system  under
     negative pressure,  which pulls  the sample  through the  sampling train and
     into the Tedlar® bag.  The  sampling train vacuum is monitored with a vacuum
     gauge  installed in-line between the vacuum line and the coarse adjustment
     valve mounted in the control console.

          4.1.3    Sample flow rate is regulated by adjusting the  coarse and fine
     valves on the control  console.   The  coarse adjustment  valve controls the
     sample inlet  volume  and  rate and isolates  the vacuum  line, vacuum gauge,
     and  sample  train  from the pump and other  console components during leak
     checks.   Sample  volume  is  measured  with a  calibrated  dry gas  meter
     contained  in the  control  console.    Sampling  rate  is monitored  by  a
     rotometer, contained in the control console, and is installed  on the outlet
     side of the dry gas meter.
                                   0040 - 3                         Revision 0
                                                                  January 1995

-------
     4.1.4   The  source,  probe,  filter,  and  condenser temperatures  are
monitored by Type  J or K thermocouples using the temperature readout in the
control console.  Probe  heater temperature is regulated by the  temperature
controller provided in the control  console (Figure 5).

     4.1.5   The velocity pressure  and temperature of the source gases are
measured using  a  standard  or  S-type pitot tube connected to  a  manometer
with 1/4-in. OD x  1/8-in.  ID tubing,  in accordance with EPA Method 2.   The
source velocity pressure and temperature must be monitored during sampling
and the sampling  rate adjusted proportionally to changes  in the  flue gas
velocity (Sec. 7.5.1.2).

4.2  Sample train components

     4.2.1   Probe assembly - The  probe assembly consists of  a  length of
heated and insulated borosilicate glass tube  inside  a  length of stainless
steel tubing.  The probe temperature  shall be maintained between 130°C and
140*C  in  order to prevent damage to Teflon®  lines   and  to  facilitate
efficient cooling of the  gases  in the condenser.  The stainless steel
sheath must be cooled with  water when the source temperature approaches or
exceeds 140°C.

     4.2.2   Particulate filter -  Particulate matter  from the sample gas
stream exiting  the probe  is collected on  a quartz filter substrate  in  a
heated 47-mm Teflon® or glass filter  holder.  Use clean  filters  in order
to prevent  sample contamination.   The particulate  matter itself is  not
analyzed or archived.   However, removal of particulate matter provides  a
cleaner  sample for  analysis.    All   connections  between the probe  and
particulate filter  shall  be heated  to maintain  the  temperature  between
130'C  and  140°C so that compounds remain  in  the  volatile phase.  Heat-
wrapped Teflon® unions with stainless steel  nuts  and Teflon®  ferrules are
recommended for all heated connections.

     4.2.3   Isolation  valves  -  A typical  isolation   valve  is  shown  in
Figure 3.  The  isolation valves shall  be  constructed  of  Teflon® or glass
with  Teflon® stopcocks to provide  gas-tight  seals without  the use  of
sealing greases.   The  probe  and  bag  isolation valves are of  identical
design  and  materials   and  are  therefore interchangeable.    The probe
isolation valve provides for the attachment of a charcoal  or similar purge
trap to allow  filtered  ambient air to  enter the train  when returning the
train to ambient pressure  after leak  checks.  This valve directly connects
the probe and filter assembly to the  condenser inlet and must be heated to
between 130"C  and 140°C.   The bag  isolation  valve  allows the bag to be
opened for sampling or evacuation and isolated and  sealed for  leak checks
or system purges.

     4.2.4   Condenser  -  Use  a jacketed,  water-cooled,  coil-type glass
condenser with a  volume of at least 125 ml.   The  condenser  shall  have
sufficient capacity to  maintain the  temperature of  the  sample gas stream
between 20'C and 4*C to ensure  proper removal and collection of condensable
moisture  in the  effluent gas  sample.   The  cooled   sample  gas  stream
temperature  should  not  exceed   ambient   temperature.    All  condenser
connections must form a  leak-free,  vacuum-tight seal  without using sealing
                              0040 -  4                         Revision 0
                                                             January 1995

-------
greases.  Stainless  steel  fittings  are not  permitted,  and  Teflon® unions
or washers with screw caps are recommended.

     4.2.5   Condensate  trap  -  A glass Erlenmeyer distilling  flask  with
threaded  screw  cap connections, Teflon® seals,  and  a minimum  volume  of
125 ml may be used to trap condensate.  All  connections  on the condenser
and trap shall  be sized  to accept 1/4-in. OD x 1/8-in. ID Teflon® or glass
fittings.  The  stem from the condenser must be positioned to within 0.5-in
from the bottom of the condensate trap.

     4.2.6   Sample  transfer  lines  and connection fittings -  All  sample
transfer  lines  connecting components  shall  be  less  than  5 ft  long  and
constructed of 1/4-in. OD x 1/8-in. ID Teflon® tubing  or glass.  All sample
lines upstream of the condenser  and  condensate trap must be heated and the
temperature maintained between 130°C and 140°C.   Use  Teflon® fittings for
connections between various train components to provide leak-free, vacuum-
tight connections without the  use of sealing grease.  New tubing should be
used  for  each  separate  test   series  or  condition  to  prevent  cross
contamination of sample compounds.

     4.2.7   Tedlar® storage  bag -  Choose  a bag  size  according  to  the
guidelines provided in  Sec. 7.2.4.   In order to minimize wall effects, the
sample volume must fill  at least 80% of the bag capacity.  The  recommended
size range for  bags  is  25 L  to  35 L.  Small bags  (<  25  L)  are easier to
store  and  transport but  may  have  insufficient  volume  for  proportional
sampling.  In addition,  accurate volumetric  measurement is  difficult  with
smaller bags.   Large bags  (>  50  L)  lack portability,  but may  be required
under  certain  conditions, such   as  during  proportional  sampling  and  for
sampling sources requiring high  sample rates.

     4.2.8   Evacuated container (bag container) - Use any rigid, air-tight
metal  or  plastic  (e.g.,  PVC®/Polyethylene®/Nalgene®)  drums  or  glass
containers  to  house the Tedlar®   bag  during  sampling,   storage,   and
transport. The  container must  be  constructed so  that  it  can  easily  be
assembled and disassembled (for bag removal).  The container must be  able
to hold a negative pressure of at least 10 in. H20.  The bag container must
be at least 20% smaller than  the Tedlar® bag being used but must be large
enough to hold  the volume  of  sample  required  (e.g.,  for  a  sample size of
20 L, a 25-L  Tedlar®  bag inside a 20-L container provides sufficient volume
without danger of overinflating the bag).

     Containers must not have  staples, sharp  edges,  or metal closures which
might damage bags. The  container should also be constructed of a material
that shields the sample from  exposure to sunlight  to  protect  the bag and
its contents from ultra-violet  light.   A viewing  port or  other means of
observing the flexible  bag during sampling  is desirable.   During storage
and transport,  the viewing port shall be covered with opaque material.

     4.2.9   Vacuum  lines  - Use  Tygon®, Poly®, Nylon®,  or  similar tubing
capable  of  maintaining  at least  10-in.  H20 negative  pressure  without
collapse as vacuum lines.  Tubing should be  1/4-in.  OD x 1/8-in.  ID  size
to  minimize volume  and   ensure  compatibility  of  connection  fittings
                              0040 -  5                          Revision  0
                                                             January  1995

-------
throughout the train.   Stainless steel fittings and valves may be  used  for
vacuum line connections but may not be used in the sampling line.

     4.2.10  Control console (meter system) - The metering system  required
for  this  method  is   readily  available  in  the  form  of  the   control
console/meter  box  (e.g.,  Nutech  Model 280.01B)  from  a Volatile  Organic
Sampling Train  (VOST,  Method 0030),  and  shall  consist of the  components
pictured in Figure 5.

             4.2.10.1  Vacuum  gauge   (meter  pressure)  -  Use a  direct
     reading, mechanical vacuum gauge capable of measuring pressures of at
     least 15  in.  Hg  with 1-in.  or smaller increments to monitor system
     vacuum during sampling and leak checking  the bag,  the container,  and
     the sampling train.

             4.2.10.2  Sample  flow rate adjustment valves  -  Coarse  and
     fine  adjustment  valves are  provided.  The  coarse adjustment  valve
     controls  volume  and rate of  sample flow and  isolates the  control
     console from the  sampling train  and vacuum  line during  leak checks.
     The fine  adjustment valve controls sample  rate  and system  vacuum.
     Closing the valve (clockwise) increases train vacuum and sample flow
     rate. Opening the valve (counterclockwise) decreases train vacuum  and
     sample flow rate.

             4.2.10.3  Pump  - Use a leak-free diaphragm pump or equivalent
     that is  capable of pulling and maintaining a  vacuum of at least 15  in.
     Hg and a  flow rate of  at least 1 liter per minute (Lpm).

             4.2.10.4  Calibrated  dry  gas  meter  -  The  control   console
     contains  a calibrated dry gas meter  (Singer  Model  802/American Meter
     Model 602  or  equivalent) capable of reading  1 L  per  revolution with
     0.1-L increments, and provides accurate measurement  of the volume of
     the sample collected.

             4.2.10.5  Flow  meter  -  Use a rotometer with  a glass  tube  and
     a glass, Teflon®, or sapphire float  ball  of  suitable range (0-5 Lpm)
     to measure the sample flow rate.  The flow meter shall be accurate to
     within 5% over the  selected range.  A range  of ±  25% of the desired
     sampling rate is  suggested to  ensure greater accuracy of readings  and
     a better  range for  adjustment  of the  sampling rate  (proportional  to
     the source gas stream  velocity).  The rotometer  is  installed at  the
     outlet of  the dry gas meter in the console.

             4.2.10.6  Thermocouples and temperature read-out device -  Use
     a sufficient number and length of type J or K thermocouples.   The  10-
     channel   (1 to 4  remote;  5  dry gas meter,  6 to   10  spares)  digital
     thermocouple  read-out  provided in  the control console  displays  the
     source,  probe, filter, and condenser temperatures.

             4.2.10.7  Heat   controller   -  Use   a  rheostat  or   digital
     temperature controller (e.g.,  Fuji  PYZ4  or  equivalent)  to  regulate
     probe heat temperatures.
                              0040 - 6                         Revision 0
                                                             January 1995

-------
           4.2.11   Pitot tube probe - A standard or  S-type  pitot tube must be
      used  for  pretest and post-test velocity traverses and  to monitor flow so
      that  the  sampling rate can be regulated proportionally to the source gas
      velocity  throughout  the length of  the  sampling  run.

           4.2.12   Pressure gauge (manometer) - Use a water- or oil-filled U-tube
      or  inclined manometer  capable of  measuring to  at least  10  in. H20 and
      accurate  to within 0.1  in. H20 for  monitoring and  measuring  the source gas
      velocity.

           4.2.13   Barometer -  Use an aneroid  or  other barometer  capable of
      measuring atmospheric pressure to  within 0.1  in.  Hg of actual barometric
      pressure.

           4.2.14   Charcoal  absorbent traps  -  Use   charcoal  traps to  absorb
      organic  compounds in the  atmosphere at the site.  One  charcoal trap is
      attached  to  the probe  isolation  valve  and  filters  incoming  air  when
      releasing vacuum to prevent contamination of the train  during leak checks.
      A second  charcoal trap is located  in the vacuum line and filters any gas
      exiting  the sample  train to  protect  sampling  personnel in  case  of bag
      rupture.  Any readily available,  ready made charcoal tube similar  to a VOST
      tube  may  be  used.

           4.2.15   Stopwatch -  Use  any stopwatch  capable of measuring  1 second,
      to  time sample  collection.


5.0   REAGENTS

      5.1   Reagent grade chemicals  shall  be used in all  tests.  Unless  otherwise
indicated, it is intended that all  reagents  shall conform to the  specifications
of the Committee on Analytical  Reagents  of the  American Chemical  Society, where
such  specifications  are  available.  Other grades may  be  used,  provided it is
first ascertained that the reagent is of sufficiently high purity to permit its
use without lessening  the  accuracy  of the determination.

      5.2   Water - Water used for sample train preparation shall be distilled and
deionized.   Water  used for  rinses  during  recovery  of  condensate  shall  be
prepurged high  performance liquid chromatography (HPLC)-grade.  Clean,  clear tap
water may be used as condenser cooling water.

      5.3   Nitric  acid, HN03  (10%)  -   reagent grade.

      5.4   Charcoal  - SKC  petroleum-base charcoal, or equivalent.  A mesh size
of 6-14 is recommended.  New or reused charcoal may be  used  for each run series
or test condition.  Reused charcoal  must be reconditioned using the same criteria
specified in VOST (Method  0030).

      5.5   Methanol  - Spectrometric-grade, or equivalent.

      5.6   Field spiking standards  - Appropriate gas cylinders containing the
target components of interest in known concentrations  (highest  purity available)
for field spiking.


                                   0040 -  7                          Revision 0
                                                                  January 1995

-------
6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

     6.1   Pretest  preparation

           6.1.1    Glassware  -  Before sampling, prepare the glass components of
     the  train by cleaning  with  non-ionic  detergent (e.g., Alconox)  and hot
     water in an  ultrasonic  bath.   Rinse  each component  three times  with
     distilled, deionized water, then rinse  three times with  10% HN03, followed
     by an additional three  rinses with distilled, deionized water.  Dry in an
     oven  at  130°C for  2 hours.

           6.1.2    Sample lines and bag containers - Treat  all  Teflon® lines,
     fittings, and the sample bag  containers as outlined in Sees. 6.1.1, but air
     dry these components  in an area free of  organic compounds  rather than in
     an oven.  Use clean Teflon® tubing for each test  series  or condition.  Hand
     wash  the bag  containers.

           6.1.3    Bag cleaning procedure - Ensure that all  bags are clean before
     using them  for sampling.   First,  flush  each bag three times  with  high-
     purity nitrogen  (N2; 99.998%).  Then fill  each bag with  N2 and analyze the
     bag contents  at  the highest  sensitivity setting using the same analytical
     technique that will be used for analyzing samples.  Before constructing the
     calibration curve,  analyze one  analytical system blank each day by taking
     the  gas  chromatograph  through  its analytical  program with  no  sample
     injection.  Analyze an  analytical system blank again if carryover between
     samples  is  indicated.   Other,  less stringent,  methods of  cleaning and
     analysis may  be used at  the  risk  of overlooking  important contaminants.
     An  acceptable level of contamination  will  be  a response  less  than five
     times the instrument detection  limit  or half  of the  level  of concern,
     whichever is less.   Repeat  the  nitrogen  flush  as  necessary  until  the
     acceptable level has been reached.  No  bag shall be used until it has been
     satisfactorily  cleaned.

     6.2   Sample  bag storage  and transport procedures  - To ensure sampling
integrity, perform sample recovery in a manner that prevents contamination of the
bag sample.  Protect  the bag  from sharp objects, direct  sunlight and low ambient
temperatures  (below 0°C) that could cause condensation of any of the analytes.
Store the bag samples in an area that has restricted access to prevent damage to
or tampering with  the sample  before analysis.  Analyze the  bag samples within 72
hours of sample collection  unless  it can be shown that significant (>20%) sample
degradation  does   not occur  over  a  longer period  of sample  storage.   Upon
completion of the  testing and  sample  recovery,  check all the  data  forms for
completeness  and the  sample  bags  for proper identification.   Store the bags in
rigid, opaque containers during all  sampling, storage and  transport procedures.
Ship the  bags using  ground  transportation.   Follow all  hazardous  materials
shipping procedures.

     6.3   Condensate storage  and transport procedures  - To ensure sampling
integrity, perform sample recovery in a manner that prevents contamination of the
condensate (Sec.  7.6.5). Store the condensate in 40-mL vials with no headspace.
Place the vials in  ice or in  a refrigerated container  at 4°C  (± 2°C) immediately
following  recovery and  during  transport  for analysis.   In addition,  store the
vials in an area that has  restricted access to prevent damage  to or tampering


                                   0040 - 8                         Revision 0
                                                                  January 1995

-------
with the  sample  before analysis.   Upon  completion  of the testing  and sample
recovery, check all the data forms for completeness  and the condensate samples
for  proper  identification.     Ship   the   condensate  samples  using  ground
transportation.  Follow all hazardous materials shipping procedures.

     6.4  The time lapse between sampling and analysis shall not exceed  72 hours
unless it can be justified by specific sample matrix stability data that meets
the criteria of Sec.  1.5.3.  Stability in a  Tedlar® bag shall be demonstrated by
spiking analytes  into  inert gas in the  laboratory  and into stack  gas in the
field.   The  spiking level  must be at least  at the  level found in the samples of
the emissions matrix obtained during the pre-site  survey.  Compound recovery in
both laboratory and field  studies must be > 80% after 72 hours for consideration
of applicability.


7.0  PROCEDURE

     The  overall  sampling procedure involves a pretest survey of the source to
establish sampling parameters,  a series of pretest checks of the sampling system
and the source conditions,  and  the  actual  sample  collection.   These steps are
described in Sees.  7.1 - 7.5.   Following  the actual  sample  collection step,
sampling data are recorded and a post-test  leak check is performed (Sec. 7.6).
As noted in  Sec. 1.0, this method does not  include sample analysis procedures,
but general   guidelines  for sample analysis are  given in  Sec. 7.7.   Sec. 7.8
provides an  extensive set  of calculations associated with the sample collection
and analysis procedures.

     7.1  Pretest  survey

          7.1.1    Perform a pretest survey for each source to be tested.   The
     purpose  of  the  survey  is to obtain  source  information  to  select the
     appropriate  sampling  and  analysis  parameters for that source.   Potential
     interferences  may be  detected  and  resolved  during  the  survey.   When
     necessary information  about the source cannot be obtained, collection and
     analysis of  actual source  samples may be  required.

          7.1.2    The  following information must  be collected during a survey
     before  a  test can be  conducted.   The information can be  collected  from
     literature surveys and source personnel, but an actual on-site inspection
     is  recommended.   A copy of the  survey results must  be  forwarded to the
     staff performing the sample analyses.

                   7.1.2.1   Determine   whether  the   sampling   site  is  in  a
          potentially  explosive atmosphere.   If  the sample site  is located in
          an explosive  atmosphere, use other, intrinsically safe test methods.
          This method is never to be used in a potentially explosive atmosphere
          (Sec. 1.2).

                   7.1.2.2   Measure and record the  stack dimensions  on a  data
          sheet  similar to  the data  sheet  shown  in Figure  6.    Select the
          sampling site and the gaseous sampling points according to EPA Method
          1  (Reference  9) or as  specified by the  regulatory personnel.
                                   0040 - 9                         Revision 0
                                                                  January 1995

-------
             7.1.2.3   Determine  the  stack  pressure,  temperature,  and  the
     range of velocity pressures using EPA Method 2 (Reference 9).

             7.1.2.4   Determine    the    stack    gas   moisture    content
     (Sec. 7.2.3) using  EPA  Approximation  Method 4 (Reference 9) or  its
     alternatives.  Perform the determination when process  operations  are
     as they will be during final sampling.   If the process  uses  and  emits
     ambient air, use a sling psychrometer to measure the moisture content
     of the ambient air  in the area of process  air uptake.

             7.1.2.5   Select a condensate collection  system with a minimum
     volume of 50 ml.  Select a  sampling rate and  volume that will  yield
     a total condensate catch at or below 50 ml,  to allow recovery of  the
     condensate  into  volatile  organic  analysis  (VOA) vials with  minimum
     dead space.

             7.1.2.6   In accordance  with EPA Method  1,  select a  suitable
     probe liner and probe  length as  determined by the temperature  and
     dimensions of the source.   Determine the point within  the stack that
     represents  an  average flow  and  temperature  of the stack.   Mark  the
     probe at the determined distance  to provide  a reference point.   For
     sample collection, insert the probe into the  duct to the predetermined
     point  to   ensure   proper   probe  placement   and   collection  of  a
     representative sample.

             7.1.2.7   Determine  whether the source  has  a  constant   or
     variable gas flow rate.   The flow rate may be considered  constant if
     the variation over  the  sampling period  is no  more  than 20%.  If  the
     process is  constant,  use a  constant sampling  rate  (Sec. 7.5.1).   If
     the process is not constant, use proportional sampling (Sec. 7.5.2).

             7.1.2.8   Determine  approximate  levels of target compounds by
     collecting  a pretest bag  sample for analysis.  This  information is
     needed to establish  parameters for the analytical  system.

             7.1.2.9   Check  the sampling  site   to  ensure  that  adequate
     electrical  service  is available.

             7.1.2.10  Follow all guidelines  in the health and  safety plan
     for  the  test.    Use appropriate safety  equipment as  required   by
     conditions  at   the   sampling  site (e.g.,  respirator,  ear  and  eye
     protection, and a safety belt).

7.2  Pretest procedures

     7.2.1   Assemble  the train according  to  the diagram  in Figure  2.
Adjust  the  probe,   filter,  and  valve  heater   controls  to  maintain  a
temperature between 130°C and 140°C.   Circulate cooling  water  from an  ice
bath to the condenser until the temperature  is stabilized at or below  20°C.
Allow the  probe,  filter, valve,  and condenser temperatures to  stabilize
before sampling.  Mark the probe,  pitot tube,  and thermocouple  assembly
with the  proper  sampling points  as  determined   in  accordance  with  EPA
Method 1.  Before sampling,  insert the pitot tube and thermocouple  probe
into the stack,  to allow  the thermocouple readings to stabilize.

                             0040  -  10                        Revision 0
                                                            January 1995

-------
     7.2.2   Preliminary  velocity  and temperature  traverse  - While  the
probe, filter,  valve, and condenser temperatures  are stabilizing,  perform
a preliminary velocity/temperature  traverse in accordance with EPA Methods
1 and 2.   Record  the  velocity  UP)  and temperature (T,°C) at each point to
determine  a  point of  average  flow and velocity  and measure the  static
pressure at that  point.   Determine the average velocity head  (APavg)  and
range of fluctuation.

     7.2.3   Determination of  moisture  content - Determine the  moisture
content of the gas  stream being sampled before (Sec. 7.1.2.4)  or  during
actual  sampling.    For  combustion or  water  controlled  processes  (wet
electrostatic precipitators and scrubbers), obtain moisture content of the
flue  gas  during  test  conditions  from  plant personnel  or  by  direct
measurement using EPA Method 4.

     7.2.4   Criteria for  selection of sample  volume  and flow  rate -  The
flow rate should fill the  bag  to at least 80%  of  its  capacity  during  the
sampling period.   The following criteria should be met:

             7.2.4.1   Minimum  stack sampling time for each run  should be
     1 hr.   Data from less than  1 hr of sample  collection  would  be  an
     invalid test run.  Two hours of stack sampling time  is  recommended as
     optimal.
             7.2.4.2   The  minimum  sample  volume shall be at least  15  L.

             7.2.4.3   The  minimum  sample  flow  rate  shall be 250  mL/min.

             7.2.4.4   Typically, the  average sampling flow rate  is about
     0.5 L/min, which will collect  approximately  30  L of sample per hour.

             7.2.4.5   Mass emission rate determination -  Determine whether
     the final  result will  be presented on  a concentration or mass emission
     basis  before   sampling.     If   results   will   be  presented   on  a
     concentration basis,  only the concentrations of  the target  analytes
     and the stack gas moisture content need  to be measured.   If the mass
     emission rate of any compound  is  to be presented, the volumetric flow
     rate of the stack gas must also  be determined.   The volumetric flow
     rate  may  be determined  by performing  a temperature  and  velocity
     traverse in accordance with EPA  Methods 1 and  2, with actual  sample
     collection.

7.3  Leak check procedures

     7.3.1   Bag evacuation and bag leak check procedure - Before sampling,
ensure that the Tedlar® bag is fully evacuated  and leak  free.

             7.3.1.1  Assemble the sample train as  illustrated  in  Figure
     2 and  described in  Sec.  4.1.1,  ensuring that all  connections  are
     tight.

             7.3.1.2  Turn the probe  isolation valve to  position  1  and
     turn the bag isolation valve to position 1 (Figure  4).
                             0040 - 11                         Revision  0
                                                            January  1995

-------
        7.3.1.3  Disconnect the  vacuum  line from the bag container
(the quick connect has a valve to seal the  line; Figure 2) and turn
on the pump in the control  console (Figure 5).

        7.3.1.4  Open the coarse  adjustment  valve and adjust the fine
adjustment valve on the control console (Figure 5) until the  vacuum
gauge reads 5 in. Hg.

        7.3.1.5  Turn   the   bag   isolation   valve  to  position  3
(Figure 4) and open the coarse valve completely to obtain maximum flow
rate.

        7.3.1.6  Observe the dry gas meter  and rotometer as the bag
is  evacuated.    The  bag is  completely evacuated  when no  flow  is
indicated on  the dry gas  meter  and the  vacuum  rises  to  5   in.   Hg
(Figure 5).

        7.3.1.7  Allow  the rotometer float ball  to  drop  to zero.
Time and record the leak rate using one of the following procedures.

                 7.3.1.7.1    Timed leak rate (measured in  liters per
        minute)  -  Observe  the leak  rate  indicated on  the  dry gas
        meter and time for  1 min.  The leak rate must be less than  4%
        of the  sample  rate (e.g.,  0.02 Lpm for  a  sample  rate of 1
        Lpm).

                 7.3.1.7.2    Timed  pressure loss rate  (measured  in
        inches Hg drop per minute)  -   Close  both  the coarse and fine
        adjustment valves and turn off the pump.  Observe the  vacuum
        gauge and time the  pressure drop.  The leak rate must be less
        than or equal to 0.1 in.  Hg/min.

        7.3.1.8  If  all  connections  are found to be  leak tight and
the leak rate cannot  meet the  set criteria,  discard  the bag and test
another clean bag.

        7.3.1.9  Turn the  bag  isolation valve to position 1 (Figure
4) to seal the evacuated bag.

        7.3.1.10 Turn off the pump  and turn the probe isolation valve
to  position  3 (Figure 4)  allowing  the train to  return to ambient
pressure.

        7.3.1.11 Return the probe isolation  valve to position  1, seal
the  end of  the probe  and reconnect  the  vacuum  line to  the bag
container (Figures 2 and 4).

7.3.2   Pretest leak check

        7.3.2.1  Before sampling and  immediately after evacuating and
leak checking the bag, perform a pretest  leak check of the sampling
train.
                        0040 - 12                         Revision  0
                                                        January 1995

-------
        7.3.2.2   Ensure that the bag  isolation valve  is  in  position
1 (Figure 4) and the end of the probe is sealed.

        7.3.2.3   Turn the probe isolation valve to position 2 (Figure
4), turn the pump  on, and open the coarse adjustment valve (Figure 5).

        7.3.2.4   Allow  the  sample  train to evacuate and  adjust  the
fine adjustment valve to increase the vacuum to 5 in. Hg (Figure 5).

        7.3.2.5   When  the  rotometer drops to zero  and the dry  gas
meter slows to  a  stop,  time and record the leak rate  following  the
procedures outlined in Sec.  7.3.1.7.

        7.3.2.6   If the leak rate is greater than 0.1 in Hg/min or 4%
of the  sampling rate,  check all connections,  valves, and the  probe
seal for tightness.   Any leak  found must  be corrected and  the  leak
check repeated before sampling  collection  begins.

        7.3.2.7   After  completing  a satisfactory leak  check,  return
the sampling train to  ambient pressure by turning the probe isolation
valve to position 3 (Figure  4)  and  turning  off the  pump (Figure 4).

        7.3.2.8   When  the  vacuum  gauge drops  to zero,  immediately
turn the probe isolation valve  to position 1 (Figure 4).

7.3.3   Post-test leak check

        7.3.3.1   A post-test leak check must be  performed after each
bag sample is collected, before  changing the bag and container for the
next sample.

        7.3.3.2   Ensure that the bag  and probe isolation  valves  are
in  position 1  (Figure 4) and  the  pump  is turned  off when  sample
collection is completed.

        7.3.3.3   Remove the probe from the stack and seal  the  end of
the probe with  a  leak-tight seal.   Check all  connections and  train
components for looseness or breakage.  Do not tighten any connections.
Record any abnormal conditions.

        7.3.3.4   Turn the probe isolation valve to position 2 (Figure
4) and  disconnect the quick connectors on  the  bag isolation  valve
return line from the tee on  the vacuum line (Figure 2).

        7.3.3.5   Turn  on  the  pump and adjust the  fine  adjustment
valve until the train vacuum  reaches  at  least  1  in.  Hg above  the
highest vacuum attained during sample collection.  Time and record the
leak rate as previously outlined in Sec. 7.3.1.7.

        7.3.3.6   If the leak rate is less  than 4% of the  sample  rate
or 0.1 in. Hg/min., the sample  is  considered valid  (Sees. 7.3.1.7.1
and 7.3.1.7.2).
                        0040 - 13                        Revision 0
                                                       January 1995

-------
             7.3.3.7   Return   the   sample  train  to  ambient   pressure
     (Sees. 7.3.2.7 and 7.3.2.8) and disconnect the sample and vacuum lines
     from the bag and container to  prepare the train  for  the  next sample.

             7.3.3.8   If the post-test leak check proves  invalid, discard
     the invalid sample.  Attach a new Tedlar®  bag, evacuate and leak check
     the bag, and repeat the sample collection.

7.4  Preparation for sample collection

     7.4.1   Perform the pretest leak checks outlined in Sec. 7.3.

     7.4.2   Remove the seal  from the end of the probe and insert the probe
into the  stack  to  the point  of  average velocity  and  temperature  and
constant flow.

     7.4.3   Purge the sampling train  (probe,  valve, and  filter  assembly
ONLY) using the following procedures.

             7.4.3.1   Cap  the  inlet  side  of the  charcoal  purge  trap
     connected to the  probe  isolation valve tee  using  a 1/4  in.  cap  and
     plug with Teflon® ferrules for an air-tight  seal (Figure 2).

             7.4.3.2   Disconnect the vacuum line quick connect  from  the
     rigid bag container (the quick connect  has a valve  to seal the line).

             7.4.3.3   Disconnect  the  return  line connected  to  the  bag
     isolation  valve  from  the  quick connect  at the  vacuum  line  tee
     (Figure 2).

             7.4.3.4   Connect  the  purge  line from  the  probe  isolation
     valve tee to the vacuum  line tee  using the quick  connects (Figure 2).

             7.4.3.5   Ensure that  the bag isolation  valve  is in  position
     1 (Figure 4), turn on  the  pump,  and turn the  probe  isolation  valve to
     position 2  (Figure 4).

             7.4.3.6   Draw at  least eight times the sample volume of flue
     gas, or purge for at least 10 minutes, whichever is greater.

     NOTE:   A  three-way  valve may be used  in place of the  purge  quick
             connects  at the vacuum line  tee.

     7.4.4   Adjust the sample flow rate to the desired  setting and check
all  temperature  and  flow readings  during the  purge  to  ensure  proper
settings.

     7.4.5   Purge the sampling train before and between the collection of
each sample during the test run.

     7.4.6   Label each bag/container and VGA vial  clearly,  uniquely,  and
consistently with its corresponding data form  and run.  Follow appropriate
traceability requirements as defined  by the regulatory  personnel.
                             0040 - 14                         Revision 0
                                                             January 1995

-------
     7.5   Sample  Collection

     Start sample collection after the pretest leak check (Sec. 7.3.2) and the
system purge (Sec. 7.4).  Collect the sample using  proportional  rate sampling if
the pretest survey measurements (Sec.  7.1.2.7)  show that the emission flow rate
varies by more than  20% over  the sampling  period.  Otherwise, use constant rate
sampling.   Prepare for  sample  collection  for  either  method  by turning the bag
isolation valve to position 2 (Figure 4) while the pump is still  running from the
system purge.

     If  a  viewing  port has been  incorporated in  the bag  container design,
visually inspect the Tedlar® bag frequently during  the sampling run to ensure
that it is  filling properly  and that  a  sufficient sample  volume is collected.
This frequent inspection will also help prevent overfilling and  bursting the bag
during sampling.

           7.5.1   Constant rate sampling

                  7,5.1.1   Place the end  of  the  probe at a  point  within the
           duct  determined  to have  the average velocity and  temperature  and a
           constant flow rate.

                  7.5.1.2   Record  the start volume from the dry gas meter and
           begin timing  the sample period.

                  7.5.1.3   Take flue  gas velocity  and  temperature  readings
           using  either EPA  Method  2A for  smaller  ducts  (<24  inches) with  a
           remote  pi tot  tube  and thermocouple  or EPA Method  2 for larger ducts
           (>24  inches).   Utilizing  a  sample  probe  with  pitot  tubes  and
           thermocouples attached will generally ease sampling and will provide
           a direct  means  to  monitor flue gas  velocity  and  temperature  at the
           sample  probe  inlet.

                  7.5.1.4   Record  all  required  data  upon  starting,  and  at
           intervals of  no more than  5 minutes  on  the field  sampling data form
           (Figure 7).

                  7.5.1.5  Adjust  the  sample flow  rate and   sampling  train
           heating  systems to  the  correct  levels,   after  every velocity  and
           temperature reading.   The tester must closely monitor  the sample train
           and control console to ensure  that the sample  flow rate does not vary
           by more than  20% during any 5-minute period.

           7.5.2   Proportional  sampling

                  7.5.2.1   Position  the  probe  in  the center  of the  stack.

                  7.5.2.2  Record  the start volume from the  dry gas meter and
           begin timing  the sample period.

                  7.5.2.3  Monitor   the   velocity  head  during  sampling  as
           described in Sec. 4.1.5 and maintain a constant proportion between the
           sample flow rate and  the flow rate in the duct.   The flow rate  to be
           used during  sampling  (Sec. 7.2.2) is calculated using the proportional

                                  0040 -  15                         Revision 0
                                                                  January 1995

-------
     sample rate equation  in Sec. 7.8.4.  With this equation and the sample
     rate assigned to the average flow rate,  the rotometer setting can  be
     determined  after each  velocity  reading  and  the  sample  rate set
     accordingly.

             7.5.2.4   Record  all  required  data  upon  starting,  and  at
     intervals of no more than 5 minutes on the  field sampling data  form
     (Figure 7).

     7.5.3   Single-point sampling

     Collect samples from a single point within  the duct  as described  in
Sees. 7.5.1.1 and 7.5.1.2, unless multipoint  sampling has  been determined
necessary (Sec. 7.5.1.4).

     7.5.4   Multipoint sampling

     Perform multipoint integrated sampling only in  a case where  there  is
a possibility of effluent stratification. Stratification  of gases is  less
likely than of particulates.   If however, multipoint sampling  is required,
determine the  necessary  number of sample  points in accordance  with EPA
Methods 1 and 2.

7.6  Post-test procedures

     7.6.1   Record the final volume from the dry gas meter at the end  of
each sample collection period.

     7.6.2   Perform a post-test leak check as described  in Sec.   7.3.3.

     7.6.3   Inspect the  field  sampling data form (Figure 7)  and sample
identification labels for accuracy and completeness.

     7.6.4   Replace the  particulate filter after each sample.

     7.6.5   Condensate Recovery - The condensate collected during  sampling
must be  recovered separately for  each  individual  bag  sample  collected,
using the following procedures.

             7.6.5.1   Carefully  remove  the condensate trap, the condenser
     and the sample line (from the  trap  to the bag) from the sample train.
     Pour  the  contents  of  the  condensate  trap into  a  clean measuring
     cylinder.

             7.6.5.2   Rinse  the condenser, the  condensate trap  and the
     sample three times with  10 ml of HPLC grade  water and add  the  rinsings
     to the measuring cylinder containing the condensate.  Record the final
     volume of the condensate and rinse  mixture on the field sampling data
     form  (Figure 7).   High moisture  sources   (such as  those  with wet
     control devices)  may require  a 150-mL or  200-mL measuring  cylinder
     while low moisture  sources  (such  as some rotary kilns and pyrolytic
     incinerators) may require only a 100-mL size.
                             0040  -  16                         Revision 0
                                                             January 1995

-------
                   7.6.5.3   Pour the contents of the measuring cylinder into a
           20-  or 40-mL amber glass VOA vial with  a  Teflon®  septum screw cap.
           Fill the vial  until the liquid level rises above the top of the vial
           and  cap tightly.  The vial should contain zero void volume (i.e., no
           air  bubbles).  Discard any excess condensate into  a separate container
           for  storage  and  transport for proper disposal.

                   7.6.5.4   Label  each  vial by using wrap around labels.  Labels
           can  be preprinted or  can be  filled out on  site.

      7.7   Analytical Approach

      The  following  description provides general guidelines  to  the analytical
approach rather than a comprehensive analytical protocol.  The primary analytical
tool  recommended for the measurement  of volatile  organic  compounds  in source
emissions is GC/MS using fused-silica capillary GC  columns  such as described in
Method 8260.  Prescreening  of the sample by gas chromatography with either flame
ionization (GC/FID) or, for electronegative compounds, electron capture detection
(GC/ECD) is recommended because  it may not only be cost  effective, but will also
yield information regarding the complexity and concentration of the sample.  If
the smallest feasible injection  loop saturates the  analytical system, dilutions
of the sample can be made into  Tedlar®  bags using pure  N2 (99.998%) as diluent.
Calculate the  concentration  of the volatile organic compounds  in  the gaseous
emissions by using the equations (14-18) in Sec.  7.8.10.

           7.7.1   Analysis of gaseous components -  Introduce  the gases into the
      gas  chromatograph  through  the use of  a sample loop.  Use a cryogenic trap
      if sample concentration before analysis if necessary.

           For  most purposes,  electron ionization  (El)  mass spectra  will  be
      collected because  a  majority  of the  volatile  organic compounds  give
      characteristic El  spectra.  Also, El  spectra  are compatible with the NIST
      Library of  Mass Spectra and  other mass  spectral references,  which aid in
      the  identification process for other  components  in the incinerator process
      streams.

           To  clarify some  identifications,  chemical ionization  (CI)  spectra
      using  either positive  ions  or negative  ions  can  be used  to  elucidate
      molecular-weight  information  and simplify the  fragmentation  patterns of
      some compounds.   In no case, however, should  CI spectra alone be used for
      compound  identification.  For descriptions of GC conditions, MS conditions,
      internal  standard usage, and  quantitative and  quantitative identification,
      refer  to  Method 8260,

           7.7.2   Analysis of  condensates - Refer to Method 5030  to analyze
      condensate  samples by using  the  purge and trap  technique or  by direct
      aqueous  injection.   Use direct solvent injection  if  an  organic phase is
      present  distinct  from the aqueous phase.   Use dilution as  necessary to
      prevent saturation  of the  analytical  system.
                                   0040  -  17                         Revision 0
                                                                  January 1995

-------
 7.8   Calculations
      7.8.1    Carry out all calculations for determining the concentrations
 and  emission rates of the target  compounds.   Round off figures  to  three
 significant  figures after final  calculations.
      7.8.2
A
AB
Ac
AVO
AT
 Emission
**stock

D»
  eff(std|
Nomenclature
 Stack/source cross sectional  area,  m2  (ft2)
 Amount of volatile organic compound in bag (ng)
 Amount of volatile organic compound in condensate (ng)
 Amount of volatile organic compound  in analytical sample (ng)
 Total  amount of volatile organic compound (ng), AB + Ac
 Water   vapor  in  the  gas  stream,  proportion  by  volume
 (x!00=%H20)
 Type  S  pitot  tube  coefficient  (nominally  0.84 ± 0.02),
 dimensionless.
 Concentration  of  volatile  organic compound   in  emissions
 (M9/L)
 Concentration of volatile organic compound per volume sampled
 (M9/L)
 Concentration of spiking standard  in  the Tedlar® bag
 Concentration of spike standard in the stack/audit cylinder.
 Volumetric flow rate of exhaust gas,  L/min, ft3/ro-
 Pitot tube constant,
                    34.97m/sec
                                              11/2
                                  gmole
                                       ) (mmHg)
                                  (K) (mmH,0)
                   85.49 ft/sec
                                    Ib
                                  1bmole
                                        HinHg)
                                  (°R)(inH20)
                                                1/2
                              0040 - 18
                                                  Revision 0
                                                January 1995

-------
La          = Maximum  acceptable leakage  rate  for a  leak check,  either
              pretest or following a component change; equal to 0.00057  L
              (0.02 ft3/min) or 4% of the average  sampling  rate, whichever
              is less.

LDLVO,      =  Lower  detectable  amount  of  volatile organic  compound in
              entire sampling  train.

L,          =  Individual  leakage  rate   observed  during  the  leak check
              conducted  before to the  "ith" component change  (i  = 1, 2,
              3...n) L/min.

Lp         =  Leakage rate observed during the post-test leak check, L/min.

Max Massvol =  Maximum allowable  mass flow  rate (g/hr [lb/hr])  of  volatile
              organic compound emitted  from the combustion source.

Max Concvol  = Maximum  anticipated concentration  of the  volatile  organic
              compound in the  exhaust gas  stream,  g/m3 (Ib ft3).

Md         =  Stack-gas dry molecular weight, g/g-mole (Ib/lb-mole).

Mfd         =  Dry mole fraction  of the  flue gas.

M8         =  Wet molecular weight of the  flue gas.

Mw         =  Molecular weight of water,  18.0 g/g-mole (18.0 Ib/lb-mole).

Pbar        =  Barometric pressure at the sampling site, mm Hg  (in. Hg).

Pg         =  Flue gas static  pressure,  mm H20 (in. H20).

Pk         =  Specific gravity of mercury  (13.6)

P8         =  Absolute stack gas pressure,  mm Hg  (in.  Hg).

Pstd        =  Standard absolute  pressure,  760 mm  Hg (29.92 in.  Hg).

Qm         =  Average sampling rate, L/min.

Qs         =  Calculated sampling rate,  L/min.

Qsd         =  Volumetric air flow rate,  (L/min, ft3/min).

R          =  Ideal gas constant, 0.06236 mm Hg-m3/K-g-mole (21.85 in. Hg-
              ft3/cR-lb-mole).

Tm         =  Absolute average dry gas  meter temperature,  °K  (°R).

T.         =  Absolute average stack gas  temperature,  °K (°R).

Tstd        =  Standard absolute  temperature, 293°K (528°R).

                             0040 -  19                         Revision  0
                                                             January  1995

-------
V;

V, cone

V
 m(std)
U
* spike
VT
"w(std|
WF
Y
AH
AP
AP
p*
9
  avg
Analytical sample volume (ml).
Bag volume (ml).
Concentration of volatile organic compound (wt %) introduced
into the combustion process.
Anticipated concentration of  the volatile  organic compound
in the exhaust gas stream,  g/L (lb/ft3).
Total volume of liquid collected in the condensate knockout
trap.
Volume of gas sample as measured by dry gas meter, L.
Volume of gas sample measured  by dry gas meter, corrected to
standard conditions, L.
Volume of gaseous or liquid spiking standard  (ml)
Minimum dry standard volume to be collected  at  dry gas meter.
Train sample volume  (ml)
Volume  of water  vapor  in  the  gas  sample,  corrected  to
standard conditions, L (ft3).
Stack gas  velocity,  calculated by  Method  2,  Equation 2-9,
using data obtained from Method 5,  m/sec (ft/sec).
Mass flow rate of waste feed per hour, g/hr (Ib/hr).
Dry gas meter calibration factor, dimensionless.
Average pressure differential  of orifice meter, inches H20.
Actual velocity pressure, mm  (in.)  H20.
Average velocity pressure,  mm (in.) H20.
Density of water, 0.9982 g/mL (0.002201 Ib/mL).
Total sampling time, min.
Sampling time interval  from  the  beginning of a run until the
first component change, min.
Sampling  time  interval  between two   successive  component
changes, beginning with  the interval  between  the first and
second changes, min.
                              0040 - 20
                                                 Revision 0
                                               January 1995

-------
0p         =  Sampling time interval from the final (nth) component  change
              until the end of the  sampling run, min.

60         =  Second/minute conversion.

100        =  Percent conversion.


      7.8.3    Conversion  factors

      To  convert  cubic  feet  (ft3) to liter (L), multiply  by  0.02832.

      7.8.4    Proportional  sample rate  calculation

      The flow rate to be  used during sampling when the  velocity head  varies
from  the average is calculated using the following equation:

                           Q .,
                           M«   Mm
      7.8.5    Dry gas  volume -   Correct  the  sample measured  by  the dry gas
meter to standard  conditions  (20"C, 760 mm  Hg) by  using  the  following
equation:


                    T..  Pk  + AH/13.6          PK + AH/13.6
         V    = V / —   bar           = K V v   b"
          m(std)    m'  j        p          *1 m'       -r
                     1 m        std                     m

where:

K, =  0.3858 K/mm Hg for metric units,  or

K, =  17.64°R/in. Hg for English units.

      Equation 2  can be used as written, unless the leakage rate observed
during any of the mandatory leak checks (i.e.,  the post-test leak check or
leak  checks  conducted before component  changes) exceeds  La.   If Lp or L|
exceeds  La, Equation  2 must be  modified as  follows  (with the  approval of
the appropriate  regulatory personnel):

              7.8.5.1   Case I  (no component change during sampling run)

              Replace  Vm  in Equation 2  with  the expression:
                              0040  -  21                         Revision 0
                                                             January 1995

-------
              7.8.5.2  Case II   (one  or more  component changes during the
                       sampling  run)

              Replace Vm in Equation 2 with the expression:
                   V  -
                    m
                         1=1
     and substitute  only  for those  leakage  rates  (Lj or Lp)  that  exceed La

     7.8.6    Volume  of water vapor
                     V     = V  —
                      w(std)   "lc u

                                     K2V1c
                                 w   std
where:

K2 =  0.001333 m3/mi for metric units, or

K2 =  0.04707 ft3M for English units.

     7.8.7    Moisture content

                                   V
                         B
                                    w (std)
                          WS    V     + V
                                m (std)    w (std)
                                                                            i
     7.8.8
        Volumetric flow  rate  equations

        7.8.8.1   Static  pressure

                                 P
                      P  = Pn  +
                       s    Bar



        7.8.8.2   Dry molecular weight


Md = (% C02 x 0.44) +  (% 02 x  0.32)  + [(% CO  + % N2) x 0.28]



        7.8.8.3   Dry mole fraction
                              0040 - 22
                                                           Revision 0
                                                         January 1995

-------
             7.8.8.4   Wet  molecular weight

                     M  = (M. x M,.)  + (18 x B  )
                      8   x d    fd'   v      ws'


             7.8.8.5   Flue gas  velocity
                     v = k  c
                                   avg
                                         M P
                                          s  s
             7.8.8.6   Volumetric flow rate
                 DV.ff(std, - 60 Vs
fd
                                       std
                                      8 (avg)
                                                std
     7.8.9   Concentration  of  a  volatile  organic compound in the gaseous
emissions of a combustion process

             7.8.9.1   Divide  the  amount  of  volatile organic  compound
     determined through analysis by  the volume of sample introduced  into
     the analytical system to obtain  concentration of the volatile organic
     compound in the bag or  the condensate.
                                    vol
                              vol
             7.8.9.2   Multiply the concentration of the volatile organic
     compound (ng/l) by the sample volume (bag  or condensate) to determine
     the amount of the  volatile organic compound in the bag  or condensate.
             or
                               = C  , x V0
                                 vol    B
                            A  = C  , x V,
                             C   vol    Ic
             7.8.9.3   Sum the amount of volatile organic compound found  in
     all samples associated with a single train.
                            AT = AB
             7.8.9.4  Divide the total  amount found by  the volume of stack
     gas sampled  to  determine the concentration  of the volatile organic
     compound in the gaseous  emissions.
                             0040  - 23
                               Revision 0
                             January 1995

-------
                                   T
                                  -I = Cc  . .
                                  if     Emission
                                  VT
           7.8.10   Concentration  of the  spiking  standard in  the Tedlar® bag

                                     V    X C
                              r-    _  spike    stock
                               spike        y
                                          VB


           7.8.11   Recovery of the  spiking standard from the  Tedlar® bag sample


                            % Recovery = _J£L X  100
                                          spike
8.0  QUALITY CONTROL

     8.1   Quality  assurance/quality  control  requirements  before  sampling

           8.1.1    Pitot tube probe -  Before sampling, assemble and calibrate the
     pitot tube  probe  (described  in  Sec. 4.2.11)  in  accordance  with  EPA
     Method 2.   Leak  check to ± 10  in. H20.   The  pitot tube assembly must be
     leak  free  (0.00  in.  H20 in 1 minute).

           8.1.2    Pressure gauge  (manometer)  - Calibrate  the  pressure gauge
     (described in Sec. 4.2.12) in accordance with  EPA Method 2.   Leak check the
     pitot tubes,  pressure gauge, and  pitot tube  lines  simultaneously, as a
     unit,  before  the velocity  traverse.

           8.1.3    Thermocouple  and temperature read-out  device - Calibrate these
     devices  (described  in Sec. 4.2.10.6)  within  30 days  of  sampling  and in
     accordance with  EPA  Method 2.  The thermocouple and temperature  read out
     must  be accurate to  ± 1"C.

           8.1.4    Metering system  - Calibrate the dry gas meter contained in the
     control console  in accordance with the  procedures  outlined  in  Sec.  5.3 of
     EPA  Method 5.   Calibrate  the meter at  a flow rate  appropriate for the
     sampling rate used during  the test.

           8.1.5    Probe  heater -  Calibrate  the probe  heater  before  sampling
     collection following procedures outlined  in Sec.  5.5 of EPA Method 5.

           8.1.6    Barometer -  Adjust the barometer daily and before  each test
     series  to ±  0.1 in.  (25  mm)  Hg  of  the  corrected  barometric  pressure
     reported by  a National Weather  Service Station located nearby and  at the
     same  altitude above  sea level.
                                   0040 -  24                        Revision 0
                                                                  January  1995

-------
     8.2   Number  of  sampling  runs

     The  number of sampling  runs  to  be performed shall be  determined  by the
appropriate regulatory personnel.  At least two runs (two hours of stack sampling
time) are recommended for each test series  to provide minimal statistical data.

     Ensure that  all compounds on the analyte list have  been validated for this
method prior to sampling.  Perform validation as  required in accordance with EPA
Method 301 (Reference 9).

     8.3   Blanks  and  field  spikes

     Field,  trip  and  laboratory  blanks, contamination  checks and  field spiked
samples are required to monitor the  performance  of  the  sampling method and to
provide the  required  information  to take  corrective  action if  problems  are
observed in the laboratory operations or in field sampling activities.

           8.3.1    Field  blanks - Take at least one field blank sample daily and
     per source.  Collect high purity air or N2  (99.998%) from a compressed gas
     cylinder  in  the same manner as source emissions.  Draw the air or nitrogen
     gas  through  the sampling system  and  into  the  bag.  Field blank  samples
     shall consist of  the condensate and a bag  sample.   Transport and  analyze
     this  blank sample along  with  the  stack gas samples.  When the field blank
     values are greater than 20% of the stack values, flag the data.  Report the
     field blank  values with  the stack gas  results.

           8.3.2    Trip blanks -  Take  at least two  Tedlar®  bags labeled "trip
     blanks" and filled with an inert gas to the  sampling site.  These bags will
     be  treated like  any other  samples except  that they will  not  be  opened
     during  storage  at the  site.   These bags will be subsequently analyzed to
     monitor  potential  contamination  which  may  occur during  storage  and
     shipment.

           8.3.3    Laboratory  blanks - Leave two Tedlar® bags labeled "laboratory
     blanks"  in  the  laboratory using  the  method of storage that  is used  for
     field samples.   If  the field  and trip blanks contain high  concentrations
     of  contaminants  (i.e.,  greater  than five  times  the detection limit  of
     particular analyte), the laboratory blank shall  be analyzed to identify the
     source of contamination.

           8.3.4    Tedlar® bag contamination checks  - The use of  new bags  for
     each  test series is recommended.  All  bags must be  cleaned and checked for
     contamination before being used for sampling (Sec.  6.1.3).

           8.3.5    Field spike samples - Take at least one field spike sample per
     10 field samples, or a minimum number of one field spike per  test.   Spike
     the chosen bag sample with a  known mixture (gaseous or liquid) of  all  the
     target pollutants using  either  gaseous or  liquid  injection into the bag.
     Transport and analyze the spiked sample with the stack gas samples.   Report
     the  spike  sample  recoveries  with the source test  results.   The compound
     recoveries in the spiked sample must be 80 - 120%.   Use Equation 17  in Sec.
     7.8.11 to calculate  spiking compound  recovery.
                                  0040  - 25                         Revision 0
                                                                  January 1995

-------
          The  spiking  concentration  should  be  at  least  the  concentration
     anticipated  in  the  emissions  matrix.   Use Equation 16 in Sec.  7.8.10  to
     calculate the spiking concentration.  The  syringe  volume  for  the  gaseous
     injection should not exceed 200 ml to minimize leakage through the septum
     after  injection.    For  liquid injections, the  volume injected must  not
     exceed  1 ml  to  ensure  complete volatilization.   The  final  volume  of  the
     spiked  gas must not exceed  1% of the  total sample  volume.   Use the ideal
     gas equation to  calculate the  volume of gas generated by a liquid injection
     into the bag.

                  8.3.5.1  Obtain   spiking   stock   that   is   sufficiently
          concentrated to spike  a  Tedlar® sample without  exceeding  1% volume
          limit.  Select  appropriate analyzes, analyte homologs, or isotopically
          labeled  analogs  in  cylinders  or  SUMMA®  canisters   for  gaseous
          injections  or  neat   liquids   or   methanol   solutions  for   liquid
          injections.

                  8.3.5.2   Install an   injection  port  that  consists  of  a
          Swagelok®  tee  fitting  with a septum,  in  the  sample  line  just before
          the 1/4-in. quick  connector  on the Tedlar® bag  (Figure  2).   Locate
          this port  as close  to  the bag  as possible  to  minimize  wall effects.
          Use a new  septum for each sampling run that involves spiking.

                  8.3.5.3   Perform a  leak test as described  in  Sec. 7.3 with
          the injection  port  in  line.

                  8.3.5.4   Start sampling the stack  as  described in Sees.  7.4
          and 7.5.

                  8.3.5.5   In preparation for injection,  clean the syringe by
          flushing three times with an inert gas (high purity N2, 99.998%)  for
          gaseous  injections, or  with methanol  for  liquid injections.  Then
          flush  the  syringe  three times with  the  gaseous or liquid  spiking
          standard.

                  8.3.5.6   After half an hour of sample  collection,  take up  the
          desired volume of  the  spiking  standard  into  the  syringe  (for gases,
          allow the standard to equilibrate to  atmospheric  pressure) and inject
          it through the septum  into the bag without interrupting the sampling
          procedure.  All apparatus upstream of the bag should be under slight
          negative pressure.

     8.4  Performance audits  should be  conducted to evaluate quantitatively  the
quality of  data  produced by  the total  measurement system  (sample  collection,
sample analysis,  and  data processing).  Accuracy (% recovery) must be 50 - 150%.
Precision  (% relative  standard deviation)  must  be less  than   or equal   to
50 percent.

     8.5  Evaluation of analytical  procedures for a selected series of compounds
shall include the sample preparation procedures and each associated analytical
determination.  Challenge the analytical procedures by spiking the  test compounds
at appropriate levels carried through  the procedures.
                                   0040  -  26                         Revision 0
                                                                  January 1995

-------
     8.6   Determine  the  overall method detection limits (lower and upper) on a
compound-by-compound  basis   according   to  the  40 CFR   Part  136b   for  the
determination of the  detection limit.  Different compounds may exhibit different
collection efficiencies as well  as instrumental minimum detection limit.

     8.7   Determine  the  overall  method  precision  and bias (in accordance with
the EPA Method  301)  on  a compound-by-compound  basis at  a  given concentration
level.    Include in the method precision  value a combined variability  due to
sampling  and  instrumental  analysis.   The method  bias  is dependent  upon the
collection efficiency of the train components.


9.0  METHOD PERFORMANCE

     No evaluation or validation data are currently available for this method.


10.0 REFERENCES

1.   Howe,  G.B.,  Pate,  B.A.,  and  Jayanty,  R.K.M.,  "Stability of  Volatile
     Principal Organic Hazardous  Constituents  (POHCs) in Tedlar® Bags", Research
     Triangle Institute  Report to  the EPA, Contract  No. 68-02-4550,  1991.

2.   Andino,  J.M.,  and Butler,  J.W.,  "A  Study of the Stability of Methanol-
     Fueled Vehicle  Emissions in Tedlar® Bags", Environ.  Sci.  Technol. 1991,
     25(9), 1644-1646.

3.   Posner,  J.C.,  and Woodfin,  W.J.,  "Sampling  with Gas  Bags I:   Loses of
     Analyte with Time", Appendix  L  Industrial  Hygiene, 1986, (4),  163-168.

4.   Seila,  R.L.,  Lonneman,  W.A.,  and  Meeks, S.A., "Evaluation of Polyvinyl
     Fluoride as a Container Material  for Air Pollution  Samples",  J. Environ.
     Sci.  Health., 1976, 2,  121-130.

5.   U.S.  Environmental   Protection  Agency,  Hazardous   Waste  Incineration
     Measurement Guidance Manual,  Volume III of the  Hazardous Waste Incineration
     Guidance Series, EPA/625/6-89/021, p. 5.

6.   U.S.  Environmental  Protection Agency, Method 301, "Protocol for the Field
     Validation  of Emission Concentrations from Stationary Sources", EPA 450/4-
     90-015, February 1991.

7.   U.S.  Environmental  Protection  Agency,  40  CFR  Part  136, Appendix  B,
     "Definition and Procedure  for the Determination of  the  Method Detection
     Limit".

8.   Kanniganti,  R.,  Moreno,  R.L.,  and  Bursey,   J.T.,  Radian  Corporation,
     Research Triangle Park,  North Carolina, "Method 0040: Sampling of Principal
     Organic Hazardous Constituents from Combustion Sources Using  Tedlar® Bags",
     EPA  Contract No. 68-D1-0010.

9.   U.S.  Environmental  Protection  Agency,   40   CFR  Part  60,  Appendix  A,
     Methods 1,  2, 3, 4, 5,  18 and  25.


                                   0040 -  27                         Revision 0
                                                                  January 1995

-------
                                   TABLE 1

PROBLEMS THAT CAN INVALIDATE TEDLAR® BAG SAMPLING DATA AND SUGGESTED REMEDIES
                Problem
                Remedy
 1.    Condensation of the gases or
      water vapor in the bag


 2.    Leaks developing in the bag
      during testing, transport,
      and/or analysis
 3.    Hydrocarbon contamination
Sample below the condensation point
of the analytes; lower the
temperature in the condensate trap.

Use double sealed bags; perform
additional sampling runs; protect the
bags from sharp objects by sampling
and shipping in rigid, opaque
containers; ship the bags in the same
containers used during sampling.

Minimize exposure of the bag to heat
and direct light, by sampling and
shipping in rigid, opaque containers;
purge the bags with ultrapure N2 in
the laboratory and establish through
analysis that the hydrocarbon levels
are acceptable; use the bags only
once.
                                  0040  - 28
                            Revision 0
                          January 1995

-------
Q.


<
o   j2
00
                                                                      o in
                                                                        w
                                                                      c o>
                                                                      O -H


                                                                      Sf
                                                                      > to
                                                                      v -3
                                                                      a: c
                                                                        ro
                                                                      cn
                                                                      o
                                                                      o

-------
                                    FIGURE 2
                            ISOLATION VALVE DESIGN
                                        3-Way Glass Valve
                                       With Teflon Stopcock
                                                          r
In
            Valve Stem Reduced
              From 5/16" to 1/4"
                                                                                Out
             *
             1"-
             (All 3 Stems Of Equal
             Length, Size, & Shape)
    Heat Wrap Valve
And Heat To 13(3-140° C
                                        Out
                                     0040  -  30
                           Revision 0
                         January  1995

-------
                                      FIGURE  3
                                  VALVE OPERATION
        Isolation Position
          (Post Test)
Probe Isolation Valve Positions

     @ Sample Position         Q) Vacuum Purge Position
System Purge Position (Pre Test)   (Release System Pressure
                                  After Leak Checks)
    u) Isolation Position
System Purge Position (Pre Test)
Leak Check Position (Post Test)
   Bag Isolation Valve Positions

      @ Sample Position
) Bag Evacuation Position
 Bag Leak Check Position
       (Pre Test)
                                      0040 -  31
                                                   Revision 0
                                                 January  1995

-------
                            FIGURE  4
                  DIAGRAM OF A CONTROL CONSOLE
                            Front View
(D
    1234
                                                        LPM
                         Schematic Diagram
                                                                 LOUT
                             0040 - 32
  Revision 0
January  1995

-------
                         FIGURE 4  (Continued)
                       CONTROL CONSOLE  COMPONENTS
 1.   1/4 in. S.S. Quick Connect  - Vacuum line inlet  from sample train (to bag
      container).

 2.   Amphenol Receptacle - provides power through umbilical  to probe heat &
      water  pump.

 3.   Thermocouple Receptacles - 4 thermocouple inlets for:
      a.  Stack Temperature
      b.  Probe Temperature
      c.  Condenser Temperature
      d.  Ambient Temperature

 4.   110 VAC  Receptacle - auxiliary power for isolation valve heat.

 5.   Vacuum Gauge - 0-30 in. Hg.

 6.   Heat Controller

 7.   Digital  Thermocouple  Read Out  -  10  channel  (displays  temperature
      readings during sampling)
         (1-4  remote as listed above)
         (5  dry  gas  meter  temperature)
         (6-10 spares)

 8.   Timer  (optional)

 9.   Power  Switches -  control (on/off)

      a.  Main  power  -  with  separate  switches for  each.
      b.  Sample  pump
      c.  Water pump
      d.  Timer

10.   Meter  pressure Gauge -  (inches water column)

11.   Fine Adjustment (Bypass) Valve

12.   Coarse Adjustment  (on/off) Valve

13.   Dry Gas Meter

14.   Rotometer  (Flow Meter)

15.   Charcoal  Trap  (Optional)
                               0040 - 33                        Revision 0
                                                              January 1995

-------
                                   FIGURE 5
                           PRETEST SURVEY DATA FORM
I.     Name of Company
      Address

      Contacts
      Phone Numbers
                              Date
      Process to be sampled
      Duct or vent to be sampled
II.    Process description
      Raw material
      Products
      Operating cycle
            Check:  Batch
      Timing of batch or cycle
      Best time to test
Continuous
Cyclic
III.   Sampling site
            Description
            Site description  _
            Duct shape and size
            Materials
            Wall  thickness
            Upstream distance
            Downstream distance
            Size of port  	
      inches
      inches
                                    inches
          diameter
          diameter
                                  0040  - 34
                                Revision 0
                              January 1995

-------
                             FIGURE  5   (Continued)
      Temperature
      Velocity  _
      Static pressure
      Moisture content
      Particulate content
      Gaseous components
            N,
            0
             2
            CO
            C02
            so,
°c
01
h

01
/o
      Hydrocarbon components
     inches H20
Data Source
Data Source
Data Source
Data Source
Data Source

Hydrocarbons
                                                           ppm
                                                           ppm
                                                           ppm
                                                           ppm
                                                           ppm
                                                           ppm
ppm
ppm
ppm
ppm
ppm
B.     Sampling considerations
      Location to set up GC
      Power available at duct
      Plant entry requirements
      Security agreements
      Potential  problems
      Site diagrams (Attach additional  sheets if required)
                                   0040  - 35
                                       Revision 0
                                     January 1995

-------
             CD
                                 U
                                       Si
o in
   CT>
 C CTi
 o ^-i

 01  >>
•r-  i-
 >  fO
 0)  3
Q£  C
    re
   •-3
U





s
ai
o
u_

<
1 —
^£
^D i""**

UJ CD
rv 21
~~^ »«<
CD —1
1—1 Q-
i i ^"
^£
GO
E
rO
C
>^
"C3
^_x

. .
E
CU
-t-i
00
>.J
oo

C
o
• r-
4->
3
i—
-r—
Q

CD
c:
• r—
4->
4->
cu
t/1
3
o
f—
4-

c*
O
• t—
00
00
.f—
E
CU




CD
C
• ^
•t-J
4->
O)

2
O
r—
4-

•(->
C
0)
3
r—
.,—.
•^


(/>
• ^
+->
fO
-t->
00
s^^"

E
CU
4^J
OO
>"j
00

f^"
0
•r-
-t->
3
r—
• »^
o

CD
C
•^
-t-J
-(— I
01
00
2
o
r—
<4-

C
O
•r»
00
00
• r—
E
CU









\x
(J
0)

{_)

SX
(^
CU
— 1

, —
rO
C
• r—
u_
U

rO
O
O
r—.

-f->
C
• ^
o
a.

CD
c
• r—
, 	
a.
E
re
GO
CD
E
3
r—
o
>

a>

re
00
c
ai
•73
c
0
o

,_
re
+J
o
t—










CU
Nl
•r-
oo

,_
re
.r—
>

<£.
o
>








i^
CU
J3
E
3
C

,— —
re
-r—
>

^J
o
>





ai
E
3

O
>

CD
re
-Q

©
s_
re
p^
T3
CU
1—







CU
£
rs
r-~
O
>

t-
(D
C
•r-
re
[ ^
c
o
CJ







i.
QJ
r^
E
3
C

i_
O)
c
• r-
re
1 *
f~
o
t_J
                                                                                                                     ro
                                                                                                                     o
                                                                                                                     o
        i
                                                                     si
                                       O)









+->
c
re
,——
Q-






s-
o

re
>) S-
-I-J CU
•i- a.
o o










CU
4->
re
a




^.
a>

E
3
c

c:
3
Q:
c:
• r—
•^^

.
re

-o

NX
o
re
4->
GO
E
3
c

•x
0


CU
r—
a.
E
re
GO
a

^.^

cu
r*i
3
-(->

4->
O
-t-J
-t—
Q_


00
00
CU
i_
CL

(J
• r—
4_}
re
-l->
GO
>-
^^

c.
o
^ .^>
CU -t->
-M re
CU i-
§r*t
• r—
O r—
^- re
U_ (J
^^
c»
^n
•*.^

CU
CD
re
i.
0)
>
^^





s_
CU
i — -(->
re CU
• r— E
+-> s
•i- O
C i —
I— I <4—







CD

• r—
-(->
+->
CU
00


U
re
^)
00

CU
CD
re
s_
cu
>
^^



CU
s_
3
^J
re
s_
cu
a.
E
CU
^J






o
• r—
S- CU
->-> S-
CU 3
E oo
O oo
S- CU
rO S-
QQ Q.

-------
T3
 Ol
 c
 o
o
to


I/I
cn
Temperature Reac

01 ^~.
01 o
0 X
r— O
U. CD
o
0
Q. 01
CO -r-
OO —1
01
-O — ^
0 0
i- o
Q. --^
u . — .
rO (_>
GO • — -
s- —^
 •(->
r— QJ ^+—
ITS
01 . — .
-= 0
CM
(_) --^
o • ^ —
QJ -i— >

                                                                                                                                       >  fO
                                                                                                                                       O)  3
                                                                                                                                       o:  c
                                                                                                                                      CO
                                                                                                                                      o
                                                                                                                                      o

-------
                               METHOD 0040
   SAMPLING OF  PRINCIPAL  ORGANIC  HAZARDOUS  CONSTITUENTS
          FROM  COMBUSTION SOURCES  USING TEDLAR®  BAGS
Applicable Source
  and Analytes
                         Explosion Hazard Area
 Compounds with Boiling
 Points > 121 C
 Compounds Unstable in Bag
J         Use Other
*V     Sampling Methods
    Presurvey
  Measurements
          Select
              Condenser Trap Size
              Analytical Detector
              Flow Rate (Proportional vs Constant)
   Preparation
   and Set-up
          Assembly of the Sample Tram
          Preliminary Velocity and Temperature Traverse
          Determination ot Moisture Content
          Selection of Sample Volume and Flow Rate
       Bag
   Leak Check
Leak Rate > 0 1 in Hg in 1 mm
 or > 4% of the Sample Rate
          Discard Bag
        and Use Another
          Acceptable Leak Rate
  Pre-test Train
   Leak Check
Leak Rate > 0 1 in. Hg in 1 mm.
                         or > 4% of the Sample Rate
          Acceptable Leak Rate
                                Check All Connections
                                  Repeat Leak Check
   Insert Probe
 into Gas Stream
                                 0040  -  38
                                                            Revision  0
                                                         January 1995

-------
                            METHOD  0040  (Continued)
          SAMPLING OF PRINCIPAL ORGANIC HAZARDOUS  CONSTITUENTS
                FROM  COMBUSTION  SOURCES  USING TEDLAR®  BAGS
                     Insert Probe
                   into Gas Stream
                         I
                        Purge
                   Probe Assembly
 Collect Sample
Using Proportional
 Sampling Rate
                                              Steady
                                             Flow Rate
                                              Source
Collect Sample
Using Constant
Sampling Rate
                     Post-test Tram
                      Leak Check
Leak Rate > 0 1 in Hg in 1 mm.

 or > 4% of the Sample Rate
 Rag Invalid Test Data
Discard Invalid Samples
   Repeat Test Run
                           Acceptable Leak Rate
                   Sample Recovery
                    and Transport
                        I
                    Post-transport
                 Leak Check on Bags
  In-leakage > 20V.
                                         Out-leakage > 20%
 Flag Invalid Test Data   >.
Discard Invalid Samples   )
                           Acceptable Leakage
                       Analysis
                  (within 72 Hours of
                  Sample Collection)
                                     0040 -  39
                                               Revision 0
                                            January  1995

-------
                                 METHOD 0050

                  ISOKINETIC HC1/C1, EMISSION  SAMPLING TRAIN


1.0  SCOPE AND APPLICATION

      1.1  This method describes the collection of hydrogen chloride (HC1,  CAS
Registry Number 7647-01-0) and chlorine  (C12, CAS Registry Number 7782-50-5) in
stack gas emission samples from hazardous waste incinerators  and municipal waste
combustors.  The collected samples  are analyzed using Method 9057.  This method
collects the emission  sample isokinetically and is therefore particularly suited
for sampling at  sources, such as those controlled by wet scrubbers, emitting acid
particulate matter (e.g., HC1  dissolved in  water  droplets).  A midget impinger
train sampling method designed for  sampling  sources of  HC1/C12 emissions not in
particulate form is presented in Method 0051.

      1.2  This method is not acceptable for demonstrating compliance with HC1
emission standards less than 20 ppm.

      1.3   This method  may also  be  used  to  collect samples  for  subsequent
determination  of particulate  emissions  (SW-846  Method  0010)  following  the
additional sampling procedures described.

2.0  SUMMARY OF METHOD

      2.1   Gaseous and particulate pollutants are withdrawn  from  an emission
source and are collected in an optional  cyclone,  on a filter, and in absorbing
solutions.  The cyclone collects any liquid droplets and is not necessary if the
source emissions do not contain liquid droplets. The Teflon mat or quartz-fiber
filter collects other particulate matter including chloride salts.   Acidic and
alkaline  absorbing  solutions  collect  gaseous  HC1   and C12,  respectively.
Following sampling of emissions containing  liquid droplets, any HC1/C12 dissolved
in the  liquid  in the cyclone and/or on  the  filter  is  vaporized  to  gas  and
ultimately collected  in the impingers by pulling Ascarite IIR conditioned ambient
air through the sampling  train.  In the acidified water absorbing solution,  the
HC1 gas is solubilized and forms chloride (CV) ions.  The C12 gas present in the
emissions has a very low  solubility in acidified water  and passes through to the
alkaline absorbing solution  where it undergoes hydrolysis to  form a proton (H+),
CV' and hypochlorous  acid (HC10).   The  (CV) ions in the separate solutions are
measured by ion  chromatography (Method 9057). If desired,  the particulate matter
recovered from the filter and the probe  is analyzed following the procedures in
SW-846 Method 0010.

      2.2   The stoichiometry of HC1 and C12 collection in the  sampling train is
as  follows:    In  the acidified  water absorbing  solution,  the  HC1  gas  is
solubilized and forms chloride ions (CT)  according to the following formula:

                        HC1  + H20  = H30+ + Cl"
                                   0050 - 1                       Revision 0
                                                                  January 1995

-------
The C12 gas present in the emissions has a very low solubility  in acidified water
and  passes  through  to  the  alkaline  absorbing  solution  where   it  undergoes
hydrolysis to form a proton (H+),  Cl",  and hypochlorous acid  (HC10) as  shown:
                        H20 + C12  = H+ + Cl" + HC10
3.0  INTERFERENCES
      3.1 Volatile materials which produce chloride ions upon dissolution during
sampling are obvious interferences in the measurement of HC1.   One interferant
for HC1  is  diatomic  chlorine  (C12)  gas  which disproportionates  to  HC1  and
hypochlorous acid  (HC10)  upon dissolution  in  water.   C12 gas exhibits  a low
solubility in water, however,  and  the use of acidic  rather  than neutral or basic
solutions for collection of hydrogen chloride gas greatly reduces the dissolution
of any chlorine present.

4.0  APPARATUS AND MATERIALS

      4.1  Sampling Train.

            4.1.1 A schematic of the sampling train used in this method  is shown
      in Figure 1.  This sampling train configuration  is adapted from EPA Method
      5 and SW-846 Method 0010  procedures,  and, as such,  the majority of the
      required   equipment  is  identical   to   that   used  in  Method   0010
      determinations.   The new components required are a glass nozzle and probe,
      a Teflon union,  a quartz-fiber or Teflon  mat filter (see Section 5.5), a
      Teflon frit, and acidic and alkaline absorbing  solutions.

            4.1.2   Construction  details  for  the  basic train components are
      provided in Section  3.4 of  EPA's  Quality Assurance  Handbook,  Volume III
      (Reference 2);  commercial  models  of  this equipment  are  also  available.
      Additionally,    the   following  subsections  identify   allowable  train
      configuration modifications.

            4.1.3  Basic operating and maintenance procedures  for the sampling
      train are also described in  Reference  2.   As  correct  usage is important in
      obtaining valid results, all users should refer to Reference 2 and adopt
      the operating and maintenance procedures outlined therein unless otherwise
      specified.  The  sampling train consists of the components detailed below.

                  4.1.3.1  Probe nozzle.  Glass with  sharp, tapered  (30° angle)
            leading edge.   The taper  shall be on the outside to  preserve a
            constant I.D.  The nozzle shall  be  buttonhook  or elbow design.  The
            nozzle should  be  coupled  to the probe  liner using  a  Teflon union.
            It is recommended that a stainless  steel nut be used on this union.
            In cases where the stack temperature exceeds 210°C (410°F), a one-
            piece glass nozzle/liner assembly must be used.  A range of nozzle
            sizes suitable for  isokinetic  sampling should be  available.   Each
            nozzle shall   be calibrated  according to the procedures outlined in
            Method 0010 sec. 9.1.


                                   0050 - 2                         Revision 0
                                                                  January 1995

-------
      4.1.3.2  Probe liner.  Borosilicate or quartz-glass tubing with
a heated  system  capable  of maintaining  a  gas temperature of  120 ±
14°C  (248 ± 25°F) at  the  exit end during  sampling.   Because the
actual  temperature at the  outlet of  the  probe  is  not usually
monitored  during  sampling,   probes  constructed  and  calibrated
according to the  procedure  in Reference 2 are considered  acceptable.
Either  borosilicate or quartz-glass probe  liners  may be used for
stack temperatures up  to about 480°C (900°F).  Quartz liners  shall
be  used  for temperatures between  480 and  900°C  (900 and 1650°F).
(The softening temperature for borosilicate  is 820°C  (1508°F), and
for quartz 1500°C  (2732°F).)   Water-cooling  of the stainless  steel
sheath will be necessary at temperatures approaching  and  exceeding
500°C.

      4.1.3.3  Pitot tube.  Type S, as described in Section 2.1 of
EPA Method 2 (Reference 1).  The pitot tube shall  be attached to the
probe to  allow constant monitoring of the stack-gas velocity.  The
impact (high-pressure) opening  plane of the pitot tube shall be even
with or above the nozzle entry  plane (see Section  3.1.1 of  Reference
2)  during  sampling.   The Type S pitot tube assembly shall  have a
known  coefficient, determined as  outlined in  Section  3.1.1  of
Reference 2.

      4.1.3.4  Differential pressure gauge.  Inclined manometer or
equivalent  device as  described  in  Section 2.2  of   EPA  Method  2
(Reference 1). One manometer shall be used for velocity-head (delta
P) readings and  the other for  orifice differential pressure (delta
H) readings.

      4.1.3.5  Cyclone (optional), glass.

      4.1.3.6  Filter holder. Borosilicate glass, with a  Teflon frit
filter support and a  sealing gasket.   The sealing gasket shall  be
constructed of Teflon  or  equivalent  materials.   The  holder design
shall provide a positive  seal against leakage at  any point  along the
filter circumference.   The holder  shall  be attached immediately to
the outlet of the cyclone.

      4.1.3.7  Filter  heating system.  Any  heating system capable of
maintaining a  temperature  of  120  ±  14°C  (248 ±  25"F)  around the
filter holder  and cyclone during  sampling.  A  temperature   gauge
capable of  measuring  temperature  to  within 3°C  (5.4°F)  shall  be
installed so that  the  temperature  around  the filter  holder can  be
regulated and monitored during sampling.

      4.1.3.8  Impinger train.   The following system shall be used
to determine the stack gas moisture content and to collect HC1 and
C12: five  or six  impingers connected in series with leak-free ground
glass  fittings or any  similar leak-free non-contaminating fittings.
The  first  impinger  shown  in  Figure  1  (knockout or  condensate
                       0050 - 3                         Revision 0
                                                      January 1995

-------
impinger) is optional and  is  recommended  as  a water knockout trap
for use under test conditions which require such a trap.  If used,
this  impinger  should be  constructed as  described  below  for  the
alkaline impingers,  but with a shortened stem,  and  should contain 50
ml of  0.05  M H2S04.   The following  two  impingers (acid impingers
which each contain 100 ml of 0.05 M H2S04) shall be  of the Greenburg-
Smith  design with  the  standard  tip  (see  Method  0010,  Section
4.1.3.8).   The  next two impingers  (alkaline  impingers  which each
contain  100  mL  of 0.1 M  NaOH)  and  the  last  impinger (containing
silica  gel)  shall  be  of the  Greenburg-Smith design modified  by
replacing the tip with  a  1.3-cm  (1/2-in.)  I.D. glass tube extending
about 1.3 cm (1/2 in.)  from the bottom of the impinger (see Method
5, Section 4.1.3.8).

The condensate,  acid,  and alkaline  impingers  shall  contain known
quantities of the appropriate absorbing reagents.   The  last impinger
shall contain a known weight of  silica gel or equivalent desiccant.

      4.1.3.9   Metering  system.   The  necessary  components  are  a
vacuum  gauge,  leak-free pump,  thermometers  capable  of measuring
temperature  to  within   3°C   (5.4°F),  dry-gas  meter  capable  of
measuring volume  to  within 1%,  an orifice meter  (rate meter),  and
related equipment,  as  shown  in  Figure  1.  At  a minimum,  the pump
should be capable of 4 cfm free flow, and the dry-gas meter should
have  a  recording capacity of 0-999.9  cu  ft  with a  resolution  of
0.005 cu ft.  Other metering systems capable of maintaining sampling
rates within 10% of  isokineticity and of determining sample volumes
to within  2% may be used.   The  metering system  must  be  used  in
conjunction  with a  pitot  tube  to  enable  checks  of  isokinetic
sampling rates.

      4.1.3.10   Barometer.   Mercury, aneroid, or other barometer
capable of measuring atmospheric pressure to  within 2.5 mm Hg (0.1
in. Hg).  In  many cases, the barometric  reading may be obtained from
a nearby National Weather Service station, in  which case the station
value (which is  the absolute barometric pressure)  is requested and
an adjustment for elevation differences between the weather station
and sampling point is applied  at a rate of minus 2.5 mm Hg (0.1 in.
Hg) per 30-m (100 ft) elevation increase  (vice versa for elevation
decrease).

      4.1.3.11   Gas  density determination equipment.   Temperature
sensor and pressure gauge (as described in Sections 2.3 and 2.4 of
EPA Method 2), and gas analyzer,  if necessary (as  described in EPA
Method 3, Reference  1).   The  temperature  sensor  ideally should be
permanently attached to the pitot tube  or  sampling probe in a fixed
configuration such  that  the  tip of  the  sensor  extends  beyond the
leading  edge of  the probe  sheath and  does  not  touch  any metal.
Alternatively, the sensor may be attached just prior to use in the
field.  Note, however, that if the temperature sensor  is attached in
                       0050 - 4                         Revision 0
                                                      January 1995

-------
      the  field,   the  sensor  must  be  placed  in  an  interference-free
      arrangement with respect to the Type S  pitot tube openings  (see EPA
      Method 2, Figure 2-7).   As a second alternative,  if  the  stack gas is
      saturated, the stack temperature may be measured at a  single point
      near the center of the  stack.

            4.1.3.12  Ascarite tube  for  conditioning  ambient air.   Tube
      tightly  packed  with approximately  150 g  of fresh  8  to  20  mesh
      Ascarite IIR sodium hydroxide coated silica, or equivalent, to dry
      and remove acid gases from the ambient air used to remove moisture
      from the filter and optional  cyclone.   The  inlet and outlet ends of
      the tube should be  packed with at least 1  cm thickness of glass wool
      or filter material  suitable to  prevent  escape  of Ascarite II fines.
      Fit one end with flexible tubing, etc.  to allow connection to probe
      nozzle.

4.2  Sample Recovery.

      4.2.1  Probe liner.  Probe and nozzle brushes;  nylon bristle brushes
with stainless steel  wire  handles are required.  The  probe  brush shall
have extensions of stainless  steel, Teflon, or  inert material at least as
long as the probe.  The brushes shall be properly sized  and shaped to brush
out the probe liner and the probe nozzle.

      4.2.2  Wash  bottles.  Two.  Polyethylene or glass, 500 ml or larger.

      4.2.3  Glass sample storage containers.  Glass,  500-  or 1,000-mL.
Screw-cap liners shall be Teflon  and  constructed so as  to  be leak-free.
Narrow-mouth glass bottles have been found to  exhibit less tendency toward
leakage.

      4.2.4  Petri  dishes.  Glass  or  plastic  sealed around the circumfer-
ence with Teflon tape, for storage and transport of filter samples.

      4.2.5   Graduated  cylinder  and/or balances.   To  measure  condensed
water  to  the  nearest  1  ml   or  1 g.   Graduated  cylinders  shall  have
subdivisions  not  >2  ml.    Laboratory triple-beam  balances  capable  of
weighing to ± 0.5 g or better are required.

      4.2.6   Plastic  storage containers.   Screw-cap polypropylene  or
polyethylene containers to store silica gel.

      4.2.7  Funnel and rubber policeman.   To aid in transfer of silica gel
to container (not necessary if silica gel is weighed in field).

      4.2.8  Funnels.   Glass,  to aid  in sample recovery.
                             0050 - 5                         Revision 0
                                                            January 1995

-------
5.0  REAGENTS

      5.1  Reagent grade  chemicals shall be used in all  tests.  Unless otherwise
indicated, it is intended that all reagents shall  conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications  are  available.   Other grades may  be used,  provided it is
first ascertained that the reagent  is  of sufficiently high purity to permit its
use without lessening the accuracy of the determination.

      5.2  Reagent water.  All references to water  in the method refer to reagent
grade water  as  defined  in  Chapter  One  unless  otherwise  specified.    It is
advisable to analyze a blank sample of this water prior to sampling,  since the
reagent blank values obtained during the field sample analysis must be less than
10 percent of the sample values.

      5.3  Sulfuric acid (0.05 M),  H2S04.  Used as the HC1 absorbing reagent in
the impinger train.  To  prepare  1 L, slowly add 2.80 ml  of concentrated  H2S04 to
about 900 ml of water while  stirring,  and adjust the final  volume to 1-L using
additional water.  Shake well to mix the solution.  It  is advisable to analyze
a blank sample of this reagent prior to sampling,  since  the reagent blank values
obtained during the  field sample analysis must be less than 10  percent of the
sample values.

      5.4  Sodium hydroxide  (0.1 M), NaOH.  Used as the  C12 absorbing reagent in
the impinger train.  To  prepare 1 L, dissolve 4.00 g of solid NaOH in about 900
mL of water and  adjust the  final volume to  1-L  using additional  water.  Shake
well to mix the  solution.   It is  advisable to analyze a blank  sample  of this
reagent prior to  sampling,  since the  reagent  blank values  obtained  during the
field sample analysis must be less  than 10 percent of the sample values.

      5.5  Filter.  Quartz-fiber or Teflon mat (e.g., PallflexR TX40HI45) filter,
or equivalent.

      5.6  Silica gel.   Indicating type, 6-16 mesh.  If previously used, dry at
175°C (350°F)  for 2 hours before using.  New silica gel  may be used as received.
Alternatively, other types   of  desiccants  may be  used if  equivalence  can be
demonstrated.

      5.7  Acetone.   When  using this train for  determination  of  particulate
emissions, reagent grade acetone,  < 0.001 percent residue,  in glass bottles is
required.  Acetone from  metal containers generally has  a high residue blank and
should not be  used.  Sometimes suppliers transfer  acetone to glass bottles from
metal containers; thus,  acetone blanks shall be run prior to field use and only
acetone with low  blank values (< 0.001 percent) shall be used.   In no case shall
a  blank  value greater  than 0.001  percent  of the weight  of acetone  used be
subtracted from the  sample weight.

      5.8  Crushed ice.  Quantities ranging from 10-50 Ib may  be necessary during
a sampling run, depending on ambient air temperature.
                                   0050 - 6                         Revision 0
                                                                  January 1995

-------
      5.9  Stopcock grease.  Acetone-insoluble, heat-stable silicone grease may
be  used,  if  needed.    Silicone grease  usage  is  not  necessary  if  screw-on
connectors or Teflon sleeves on ground-glass joints are used.

6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Sample  collection  is described  in  this  method.   The analytical
procedures for  HC1  and C12 are  described  in  Method 9057 and  for particulate
matter in Method 0100.

      6.2  Samples should be stored in clearly labeled,  tightly sealed containers
between sample  recovery  and  analysis.   They may be analyzed up  to four weeks
after collection.

7.0  PROCEDURE

      7.1  Preparation for Field Test.

            7.1.1   All  sampling equipment shall be maintained and calibrated
      according to  the procedures described  in  Section 3.4.2 of EPA's Quality
      Assurance Handbook, Volume III (Reference 2).

            7.1.2   Weigh  several 200-300-g  portions of silica  gel in airtight
      containers to the  nearest 0.5-g.   Record on  each container  the total
      weight  of  the   silica  gel  plus  containers.     As  an  alternative  to
      preweighing the  silica  gel,  it may  instead  be  weighed  directly  in the
      impinger just prior to train assembly.

            7.1.3  Check filters visually against light for irregularities and
      flaws or pinhole leaks.  Label the shipping containers (glass or plastic
      Petri dishes) and keep the  filters in these containers at  all  times except
      during sampling (and weighing for particulate analysis).

            7.1.4   If a particulate determination will be conducted, desiccate
      the filters at 20 ± 5.6°C  (68 ± 10'F) and ambient pressure for at least 24
      h, and weigh  at intervals of at least  6  h to  a  constant  weight  (i.e.,
      <0.5-mg change from previous weighing), recording results to the nearest
      0.1 mg.   During each  weighing,  the filter must  not be exposed for more
      than a 2-min  period to the laboratory  atmosphere  and  relative humidity
      above 50%.  Alternatively,  the filters may be oven-dried  at  1058C  (220°F)
      for 2-3 h, desiccated for 2 h, and weighed.

      7.2  Preliminary Field Determinations.

            7.2.1  Select the sampling site and the minimum number of sampling
      points  according  to  EPA  Method  1.  Determine  the  stack  pressure,
      temperature,  and  range of velocity heads using EPA  Method 2.    It  is
      recommended  that  a leak-check  of the  pitot lines  (see EPA  Method  2,
      Section 3.1)  be  performed.  Determine the stack-gas moisture  content using
      EPA Method  4 or its  alternatives  to establish  estimates  of isokinetic


                                   0050 - 7                        Revision 0
                                                                  January 1995

-------
sampling rate settings.   Determine  the  stack gas dry molecular weight, as
described  in  EPA Method  2,  Section 3.6.   If  integrated  EPA  Method  3
(Reference 1)  sampling  is used for molecular weight  determination,  the
integrated bag sample shall be taken simultaneously with,  and for the same
total length of time as the sample run.

      7.2.2  Select a nozzle  size based on the range of velocity heads so
that it is not necessary to change  the  nozzle size to maintain isokinetic
sampling rates.  During  the  run, do not  change  the nozzle.   Ensure that
the proper differential  pressure gauge  is chosen  for the range of velocity
heads encountered (see Section 2.2 of EPA Method 2).

      7.2.3  Select  a suitable probe liner and  probe  length  so that  all
traverse points can be sampled.  For large stacks,  to reduce the length of
the probe, consider sampling  from opposite sides of the stack.

      7.2.4 The total sampling time should be two hours. Allocate the same
time to all traverse points defined by EPA Method 1.  To avoid timekeeping
errors, the length  of time  sampled at  each traverse  point  should  be an
integer or an  integer plus one-half min.  Size the condensate impinger for
the expected moisture catch or be prepared to  empty it during the run.

7.3  Preparation of Sampling  Train.

      7.3.1 Add 50  ml of 0.05 M H2S04 to the condensate impinger, if used.
Place 100 ml  of 0.05 M H2S04 in each of  the next  two  impingers.  Place 100
ml  of  0.1 M  NaOH  in  each  of the following  two  impingers.    Finally,
transfer  approximately   200-300  g  of  preweighed  silica gel  from  its
container to the  last  impinger.   More  silica  gel may be used,  but care
should be taken to  ensure that it  is not entrained and  carried out from
the impinger during sampling.  Place the silica  gel container in a clean
place for later use in the sample recovery.  Alternatively,  the weight of
the silica gel plus  impinger  may be determined  to  the nearest  0.5 g  and
recorded.

      7.3.2  Using a tweezer  or clean disposable surgical gloves, place a
labeled  (identified)  filter   (weighed,   if particulate  matter  is  to  be
determined) in the  filter holder.   Be  sure that  the  filter  is properly
centered and the gasket properly  placed to prevent the sample gas stream
from circumventing the filter.  Check the filter for tears after assembly
is completed.

      7.3.3   To  use glass liners,  install  the  selected nozzle  using  a
Viton-A 0-ring  when stack temperatures  are <260°C (500°F)  and  a  woven
glass fiber gasket when temperatures are higher.  Other connecting systems
utilizing either 316 stainless  steel or Teflon  ferrules may be used.  Mark
the probe with heat-resistant tape or by some  other method  to denote the
proper distance into the stack or duct  for each  sampling point.

      7.3.4  Set up  the train  as in  Figure 1. A minimal amount of silicone


                             0050 - 8                         Revision 0
                                                            January 1995

-------
grease may be used  on ground glass joints.  Connect temperature sensors to
the appropriate potentiometer/display unit.  Check all temperature sensors
at ambient temperature.

      7.3.5  Place crushed ice around the impingers.

      7.3.6  Turn on and set the filter and probe heating systems at the
desired  operating  temperatures.   Allow  time for  the  temperatures  to
stabilize.

7.4  Leak-check Procedures.

      7.4.1  Pretest leak-check.  A pretest leak-check is recommended, but
not required.  If the tester opts to conduct the pretest leak-check, the
following procedure shall be used.

            7.4.1.1  If a Viton  A 0-ring or other leak-free connection is
      used in assembling the probe nozzle to the probe  liner,  leak-check
      the train at  the  sampling  site by plugging  the nozzle and pulling a
      380-mm Hg (15-in. Hg) vacuum.

NOTE:  A lower vacuum may be used, provided that  it is not exceeded during
       the test.

            7.4.1.2  If a woven glass fiber gasket is used, do not connect
      the probe to the train during the leak-check.   Instead,  leak-check
      the train by first plugging  the  inlet  to  the  cyclone,  if used,  or
      the filter holder and pulling a  380-mm Hg  (15-in.  Hg)  vacuum (see
      NOTE above).  Then, connect the probe to the train and leak-check at
      about  25-mm  Hg   (1-in. Hg)  vacuum;  alternatively,  leak-check the
      probe with the rest of the sampling train  in  one  step at 380-mm Hg
      (15-in. Hg) vacuum.  Leakage rates  in  excess  of  4% of  the average
      sampling rate or 0.00057 m /min  (0.02  cfm), whichever is less, are
      unacceptable.

            7.4.1.3 The following leak-check instructions for the sampling
      train may be helpful.  Start the  pump  with bypass valve fully open
      and coarse  adjust valve  completely closed.    Partially  open the
      coarse adjust valve  and  slowly  close  the  bypass valve  until  the
      desired vacuum is reached.  Do not  reverse direction of the bypass
      valve; this will  cause water to back up  into the filter holder.  If
      the desired  vacuum  is  exceeded,  either leak-check  at  this  higher
      vacuum or end the leak-check, as  shown below,  and start over.

            7.4.1.4 When the leak-check is completed, first slowly remove
      the plug from the inlet to  the probe, cyclone,  or  filter holder and
      immediately turn  off the  vacuum pump.   This prevents the liquid  in
      the impingers from being forced backward into the  filter holder and
      silica gel  from  being entrained backward into  the fifth  impinger.
                             0050 -  9                          Revision  0
                                                            January  1995

-------
      7.4.2  Leak-checks during  sample  run.  If, during the sampling run,
a component (e.g., filter assembly or impinger) change becomes necessary
or a port  change is conducted, a  leak-check shall be conducted immediately
after the  interruption  of  sampling  and before the change  is  made.   The
leak-check  shall  be conducted  according  to  the  procedure outlined  in
Section 7.4.1, except that  it shall  be  conducted at a vacuum greater than
or equal to the maximum value recorded up to that point in the test.  If
the leakage rate is found to  be  no greater than 0.00057 m3/min (0.02 cfm)
or 4% of  the  average sampling rate  (whichever is  less),  the results are
acceptable.   If a higher leakage rate is obtained, the tester shall void
the sampling  run.   Immediately  after  a component  change  or port change,
and before sampling is reinitiated,  another leak-check similar to a pre-
test leak-check is recommended.

      7.4.3   Post-test leak-check.    A leak-check  is mandatory  at the
conclusion of each sampling run.  The leak-check shall be done using the
same procedures  as those with  the  pre-test leak-check,   except  that  it
shall be conducted at a vacuum greater  than  or equal to the maximum value
reached during the  sampling  run.  If  the  leakage  rate is found  to be  no
greater than 0.00057 m /min (0.02 cfm)  or  4% of the average sampling rate
(whichever is less), the  results are acceptable.  If a  higher leakage rate
is obtained, the tester shall void the sampling run.

7.5  Train Operation.

      7.5.1  During the sampling run, maintain  an  isokinetic sampling rate
to  within  10% of  true  isokinetic.    Maintain a  temperature  around the
filter and  (cyclone, if used) of 120 ± 14°C (248 ± 25°F).

      7.5.2   For each run,  record the data required on a  data sheet such
as the one shown in Figure 2.   Be sure to record the initial dry gas meter
reading.   Record  the dry gas  meter  readings at the beginning  and end  of
each sampling time  increment, when changes in  flow rates  are made before
and  after each leak-check,  and when  sampling is halted.    Take other
readings  required  by Figure  2 at  least once at each  sample point during
each time increment and additional readings  when significant changes (20%
variation in velocity head readings)  necessitate additional adjustments in
flow rate. Level and zero the manometer. Because the manometer level and
zero may  drift  due to  vibrations  and  temperature changes, make periodic
checks during the  traverse.

      7.5.3   Clean  the  stack  access  ports  prior to the  test  run  to
eliminate  the chance of sampling deposited material.   To  begin sampling,
remove the  nozzle  cap,  verify that  the filter and probe  heating systems
are at the specified temperature, and verify that the  pitot tube and probe
are positioned properly.  Position the nozzle at the first  traverse point,
with the tip pointing directly into the gas stream. Immediately start the
pump and adjust the flow to isokinetic  conditions  using a  calculator or a
nomograph.   Nomographs  are designed for use  when  the Type  S  pitot tube
coefficient  is  0.84 ±  0.02  and the  stack gas equivalent  density (dry


                             0050  -  10                        Revision 0
                                                            January 1995

-------
molecular weight) is equal to 29 ± 4.  If the stack gas molecular weight
and the pitot  tube  coefficient  are  outside  the  above ranges, do not use
the nomographs  unless  appropriate  steps  are taken to compensate for the
deviations  (see Reference 3).

      7.5.4    When  the  stack  is  under  significant  negative  pressure
(equivalent to  the  height of the  impinger  stem),  take care to close the
coarse adjust valve before inserting the  probe into the stack, to prevent
water from backing into the  filter holder.   If necessary, the  pump may be
turned on with  the coarse adjust valve closed.

      7.5.5  When the probe  is in  position,  block  off  the openings around
the probe and  stack  access  port to  prevent  unrepresentative dilution of
the gas stream.

      7.5.6  Traverse the stack cross section,  as required by EPA Method
1, being careful  not to  bump the  probe nozzle  into  the stack walls when
sampling near  the walls  or  when removing or inserting the probe through
the access port,  in order to minimize the change of extracting deposited
material.

      7.5.7  During the  test run,  make periodic  adjustments to keep the
temperature around the filter holder (and cyclone, if  used) at the proper
level.  Add  more ice, and, if necessary,  salt to maintain a temperature of
<20°C (68°F) at the condenser/silica gel  outlet.Also, periodically check
the level and zero of the manometer.

      7.5.8   If the pressure drop  across  the  filter  becomes too high,
making isokinetic sampling  difficult to  maintain,  it  may  be replaced in
the midst of a sample  run.   Using  another complete filter holder assembly
is recommended, rather  than  attempting to change  the filter  itself.  After
a new filter assembly  is  installed,  conduct  a leak-check.  If determined,
the total particulate  weight shall  include the summation  of all  filter
assembly catches.

      7.5.9   If  the  condensate  impinger  becomes too  full,  it  may  be
emptied, recharged  with  50  ml of 0.05 M H2S04,  and  replaced during the
sample run.   The condensate  emptied must  be saved and included  in the
measurement of the volume of moisture collected and included in the sample
for analysis.   The  additional  50 ml  of  absorbing reagent must  also  be
considered in  calculating the moisture. After the  impinger is reinstalled
in the train,  conduct a leak-check.

      7.5.10   A single train shall  be used for the  entire  sample  run,
except in cases  where  simultaneous  sampling is  required in  two  or  more
separate ducts or at two or more different locations within the same duct,
or in cases  where equipment  failure necessitates  a change  of trains.

      7.5.11  Note that when  two  or more trains are  used, separate analyses
of the  particulate  catch (if applicable)  and  the HC1  and  C12  impinger


                             0050  -  11                         Revision  0
                                                            January  1995

-------
catches from each train shall  be performed, unless identical nozzle sizes
were used on all  trains.   In that case, the particulate catch and the HC1
and C12 impinger catches from  the  individual  trains  may be combined, and
a  single  particulate  analysis and  single HC1  and  C12 analyses  of the
impinger contents may be performed.

      7.5.12  At  the  end  of  the  sample run, turn off  the coarse adjust
valve, remove the probe and   nozzle  from  the stack, turn  off  the pump,
record the final dry gas meter reading.

      7.5.13  If there  is any possibility that liquid has collected in the
glass cyclone and/or on the filter,  connect the Ascarite tube at the probe
inlet and operate the train with  the filter heating  system at 120 ± 14°C
(248 ± 25°F) at  a low flow  rate (e.g.,  AH = 1) sufficient to vaporize the
liquid and  purge any  HC1  in  the  cyclone or  on  the filter and  pull  it
through the  train into the impingers.   After 30 minutes, turn  off the
flow, remove the Ascarite tube, and examine the cyclone  and  filter for any
visible  moisture.    If moisture  is visible,  repeat   this  step  for  15
minutes.

      7.5.14  Conduct a post-test leak-check.  Also,  leak-check the pitot
lines as described in  EPA Method  2.   The lines must  pass  this leak-check
in order to validate the velocity-head data.

      7.5.15   If  the moisture  value is  available,   calculate  percent
isokineticity (see Section  7.7.10) to determine whether the run was valid
or another test  run should be  conducted.

7.6  Sample Recovery

      7.6.1   Allow the probe to  cool.  When  the  probe  can  be  handled
safely, wipe off all  the external surfaces of the tip of the probe nozzle
and place a cap  over the tip.  Do not cap the probe tip tightly while the
sampling train is cooling down because this  will create  a vacuum in the
filter holder, drawing water  from the  impingers into the  holder.

      7.6.2  Before moving  the sampling train to the cleanup site, remove
the probe, wipe  off  any  silicone grease,  and cap the  open outlet, being
careful not to lose  any condensate  that might  be present.   Wipe off any
silicone grease  and cap the filter or cyclone  inlet.  Remove the umbilical
cord from the last impinger and  cap  the impinger.   If a flexible line is
used between the  first  impinger and  the  filter holder, disconnect it at
the  filter holder  and let any  condensed  water  drain  into  the  first
impinger.  Wipe  off any silicone  grease and cap the  filter holder outlet
and the impinger inlet.  Ground glass stoppers, plastic caps, serum caps,
Teflon tape,  ParafilmR,  or aluminum  foil may be  used  to close these
openings.
                             0050  -  12                         Revision 0
                                                            January 1995

-------
      7.6.3  Transfer the probe and filter/impinger assembly to the cleanup
area.   This  area should  be clean  and  protected  from the  weather to
minimize sample contamination or loss.

      7.6.4   Save portions  of all   washing  solutions  used  for cleanup
(acetone and reagent  grade water) and the absorbing reagents (0.05 M H2S04
and 0.1 M NaOH)  as blanks.  Transfer 200 ml  of  each solution directly from
the  wash  bottle  being  used  (rinse  solutions)  or the  supply container
(absorbing reagents)  and place each in a separate,  prelabeled glass sample
container.

      7.6.5  Inspect  the train prior to  and during disassembly  and note any
abnormal conditions.

      7.6.6  Container  No.  1  (filter catch for  particulate determination).
Carefully remove  the filter  from the filter  holder and place it in its
identified  Petri  dish  container.   Use one  or more pair  of  tweezers to
handle the filter. If  it  is necessary to fold  the  filter, ensure that the
particulate cake is inside the fold.  Carefully transfer to the Petri dish
any particulate matter or  filter fibers that  adhere to  the filter holder
gasket, using a  dry  nylon bristle brush or sharp-edged blade,  or both.
Label the container and seal  with Teflon  tape  around the circumference of
the lid.

      7.6.7     Container  No.   2  (front-half  rinse  for   particulate
determination).    Taking care that dust  on  the outside of the  probe or
other  exterior  surfaces  does  not  get  into   the  sample,  quantitatively
recover particulate matter or any condensate from the  probe nozzle, probe
fitting, probe liner, and  front half of the  filter  holder by washing these
components with  acetone into a glass container.  Retain an acetone blank
and analyze with the  samples.

      7.6.8  Perform  rinses as  follows: carefully remove the probe nozzle
and clean the inside surface by  rinsing  with  acetone  from a  wash bottle
and brushing with a nylon  bristle brush.   Brush until  the rinse shows no
visible particles; then make  a  final  rinse of  the inside surface with the
acetone.  Brush and  rinse  the  inside  parts  of the Swagelok fitting with
the acetone in a similar way until  no visible  particles remain.

      7.6.9  Have  two people rinse the probe  liner with acetone by tilting
and rotating the probe  while  squirting  acetone into its  upper  end so that
all inside  surfaces  will  be  wetted with  solvent.   Let the acetone drain
from the lower end into the sample container.  A glass funnel   may be used
to aid in transferring  liquid washed to the container.

      7.6.10  Follow  the acetone rinse with  a probe brush.  Hold the probe
in  an  inclined  position  and squirt  acetone  into the  upper end  while
pushing the probe brush through  the probe with a twisting action; place a
sample container  underneath  the lower  end  of the  probe and catch  any
acetone and particulate matter that  is brushed from the probe.   Run the


                             0050 - 13                         Revision 0
                                                            January 1995

-------
brush through the probe three or more times until no visible particulate
matter is carried out with the  acetone or none remains in the probe liner
on visual  inspection.   Rinse the brush with  acetone  and quantitatively
collect these washings in the sample container.  After the brushing, make
a final acetone rinse of the probe as described  above.  Between sampling
runs, keep brushes clean and protected from contamination.

      7.6.11  Clean the inside  of the  front half  of the filter holder and
cyclone by  rubbing  the surfaces with a nylon bristle brush  and rinsing
with acetone.   Rinse  each  surface  three times,  or  more if  needed,  to
remove visible particulate.  Make a  final  rinse  of  the  brush and filter
holder.   Carefully  rinse  out  the glass cyclone and cyclone  flask  (if
applicable).   Brush  and  rinse  any particulate material  adhering  to  the
inner  surfaces  of these  components  into  the front-half rinse sample.
After all  rinses  and particulate matter have been collected in the sample
container, tighten the lid  on  the sample container  so that  acetone will
not leak out when  it  is shipped to the laboratory. Mark the height of the
fluid level to determine whether leakage occurs  during transport.   Label
the container to identify its contents.

      7.6.12   Container  No.  3  (knockout  and acid  impinger  catch  for
moisture and HC1  determination).  Disconnect  the impingers.   Measure the
liquid in  the  acid and knockout  impingers to within  ± 1 mL  by using  a
graduated cylinder or by weighing it  to within ±  0.5 g by using a balance
(if one is  available).   Record the  volume or weight  of  liquid present.
This information  is  required to  calculate the  moisture content  of  the
effluent gas.  Quantitatively transfer this liquid to a leak-free sample
storage container.   Rinse these impingers and the  connecting glassware
(and tubing,  if  used)  with water, and  add  these rinses to  the storage
container.  Seal  the container,  shake  to mix,  and label.  The fluid level
should  be marked  so that  if  any sample  is lost  during  transport,  a
correction  proportional  to  the lost  volume can  be  applied.Retain rinse
water and acidic absorbing solution  blanks  and analyze with  the samples.

      7.6.13  Container No. 4 (alkaline impinger catch for C12 and moisture
determination).  Measure and record  the  liquid in the alkaline impingers
as described in Section 7.6.12.   Quantitatively transfer this liquid to a
leak-free  sample   storage  container.   Rinse  these  two  impingers  and
connecting glassware with  water and  add these rinses to the container.
Seal the container,  shake  to mix, and label; mark  the fluid level.  Retain
alkaline absorbing solution blank and analyze with the samples.

      7.6.14   Container  No.  5   (silica gel  for  moisture  determination).
Note the color of  the  indicating  silica gel  to  determine  if  it has been
completely  spent  and make  a  notation of  its condition.   Transfer  the
silica gel from the last impinger to its original container  and seal.   A
funnel  may  make  it easier to  pour the  silica gel without spilling.   A
rubber policeman  may be used as an aid in  removing the  silica gel from the
impinger.     It  is not  necessary to  remove   the small  amount of  dust
particles that may adhere  strongly to the impinger wall.  Because the gain


                             0050  - 14                         Revision 0
                                                            January 1995

-------
      in weight is to  be  used for moisture calculations, do not use  any water or
      other liquids to transfer the  silica  gel.   If a balance is available in
      the field, weigh the container and its contents to 0.5 g or better.

            7.6.15  Prior to shipment, recheck all sample containers to ensure
      that the caps are well secured. Seal the lids of all  containers around the
      circumference with Teflon tape.  Ship all liquid samples upright and all
      particulate filters with the particulate catch facing upward.

      7.7  Calculations.   Retain  at  least one extra decimal figure beyond those
contained in the available data in intermediate calculations,  and  round off only
the final answer appropriately.

      7.7.1  Nomenclature.

            An  = Cross-sectional  area of nozzle, m2 (ft2).

            Bws = Water vapor in the gas stream,  proportion by volume.

            Ca  = Acetone blank residue concentration, mg/mg.

            Cd  = Type  S pitot  tube  coefficient  (nominally 0.84  ±  0.02),
                  dimensionless.

            cs  = Concentration of particulate matter in stack gas, dry basis,
                  corrected to standard conditions, g/dscm (g/dscf).

             I  = Percent of isokinetic sampling.

            ma  = Mass of residue of acetone after evaporation, mg.

            Mn  = Total  amount of particulate matter collected, mg.

            Md  = Stack-gas dry molecular weight, g/g-mole (Ib/lb-mole).

            Mw  =       Molecular weight of water,  18.0  g/g-mole (18.0 Ib/lb-
                        mole).

          Pbar  =  Barometric pressure at the sampling site, mm Hg (in. Hg).

            Ps  = Absolute stack-gas pressure,  mm Hg (in. Hg).

          Pstd  =  Standard absolute pressure,  760 mm Hg (29.92 in.  Hg).

             R  = Ideal  gas  constant,  0.06236 mm  Hg-m3  (K-g-mole  (21.85 in. Hg-
                  ft3/°R-lb-mole).

            Tm  =       Absolute  average dry-gas  meter temperature  (see Figure
                        2),  °K,  (°R).


                                  0050  - 15                         Revision 0
                                                                  January 1995

-------
      Ts  = Absolute  average  stack-gas  temperature  (see  Figure 2),  °K
            (°R).

    Tstd  =  Standard absolute temperature, 293°K (528°R).

      Vlc =  Total volume of liquid collected in the impingers  and silica
            gel, ml.

      Vm  =       Volume of gas sample as measured  by  dry-gas meter,  dscm
                  (dscf).

       =    Volume of gas sample measured by  the dry-gas meter,  corrected
            to standard conditions, dscm  (dscf).

Vwistd)    =   Volume of water vapor in the gas sample, corrected to standard
            conditions, scm (scf).

      Vs  = Stack-gas  velocity,  calculated  by  Method 2,  Equation  2-9,
            using data obtained  from Method 5, m/sec  (ft/sec).

      Wa  = Weight of residue in acetone  wash, mg.

      Va  = Volume of acetone blank, mL.

    Vaw   = Volume of acetone used in wash; ml.

       Y  = Dry-gas-meter calibration factor, dimensionless.

       H  = Average pressure differential across the orifice meter, mm H20
            (in. H20).

       a  = Density of acetone,  mg/jj] (see label on bottle).

       w  =       Density of water, 0.9982 g/mL (0.002201  Ib/mL).

       0  = Total sampling time, min.

    13.6  = Specific gravity of  mercury.

      60  = Sec/min.

     100  = Conversion to percent.


      7.7.2 Average dry gas meter temperature and average orifice pressure
drop.   See data sheet (Figure 2).
                             0050 - 16                         Revision  0
                                                            January  1995

-------
            7.7.3  Dry gas  volume.   Correct  the  sample measured by the dry gas
      meter to  standard  conditions  (20°C, 760 mm Hg  [68°F, 29.92  in.  Hg]) by
      using Equation  1:


Fn  T       y      _   uY      Tstd    Pbar + AH/13.6    _             Pbar+  H/13.6
tq. 1       Vm(stdl  -   VmY    	    	  ~  N VmY
                                         Pstd                        Tn
            where:      KT = 0.3858 K/mm Hg for metric  units,  or
                        K, = 17.64°R/in. Hg for English units.
            7.7.4  Volume of water  vapor.
                                n    RT
Eq- 2             Vw(8tdl = Vlc  __Pw_Jl!:td = K2 Vlc
                                Mw     Pstd
            where:      K2 = 0.001333 m3/ml for metric units, or
                        K2 = 0.04707 m3/mi for English units.

            7.7.5  Moisture content.
Eq. 3                   Bws = V«
                              V     + V
                              •mlstdl T "wlstdl


      NOTE: In saturated or water-droplet-laden gas streams, two calculations of
      the moisture content of the stack  gas  shall be made, one from the impinger
      analysis  (Equation 3)  and  a  second   from  the  assumption  of  saturated
      conditions.   The  lower of  the  two  values of  Bw  shall  be  considered
      correct.   The procedure for determining the moisture content  based upon
      assumption of saturated conditions is  given in the Note to Section 1.2 of
      Method  4.   For   the  purposes  of  this method,  the  average  stack  gas
      temperature from Figure 2 may be used to make this determination, provided
      that the accuracy  of the in-stack temperature  sensor  is  ±  1°C  (2°F).

            7.7.6  Acetone blank concentration.   For  particulate determination.
r   n                         r     "1.
Eq.  4                         Ca =   a
                                   0050 - 17                         Revision  0
                                                                   January  1995

-------
            7.7.7  Acetone wash  blank.   For participate determination.


                               Wa = Ca Vaw pa


            7.7.8  Total  particulate weight.  Determine  the total participate
      catch from  the sum of the weights obtained from Container  Nos. 1 and 2
      less the acetone  blank  (WJ .

            7.7.9  Particulate concentration.


                        cs = (0.001 g/mg)(myvm(std))


            7.7.10   Isokinetic variation.

                     7.4.10.1   Calculation  from raw data.


                     =   10° TstK3Flc + (VJTJ  (Pbar +  H/13.6)]
Eq  8
                                     600VsPsAn
            where:      K3 = 0.003454 mm Hg-m3/mL-K for metric  units,  or
                        K3 = 0.002669 in.  Hg-ft3/mL °R for English units.
7.4.10.2  Calculation for  intermediate  values.


    T =   TsVm(stdlPstd100
                               TstdVs0AnPs60(l-Bws)
                                     T V
                                     's'mlstd)
                                     PsVsAn 0 (1-BWS)
            where:      K4 = 4.320 for metric  units,  or
                        K4 = 0.09450 for  English  units.
            7.7.11   Acceptable results.   If  90% <  I  <  110%,  the results  are
      acceptable.
                                   0050 - 18                         Revision  0
                                                                   January  1995

-------
            7.7.12    Analytical   calculation  for  total  /vg  HC1   per  sample.
      Calculate as described below:
                    mHCl = (S-B) x Vs x 36.46/35.453                        (1)
            where:  mHCl = Mass of HC1 in sample, ug,
                       S = Analysis of sample, us Cl'/mL,
                       B = Analysis of reagent blank, ug Cl'/mL,
                      Vs = Volume of filtered and diluted sample, mL,
                   36.46 = Molecular weight of HC1, ug/ug-mole,  and
                   35.45 = Atomic weight of CT,  ug/ug-mole.
            7.4.13    Analytical   calculation  for  total  ug  C12  per  sample.
      Calculate as described below:
                    MC12 = (S-B) x Vs x 70.91/35.45                        (2)
            where:  Mcl2 = Mass of C12 in sample, ug,
                   70.91 = Molecular weight of C12,  ug/ug-mole, and
                   35.45 = Atomic weight of Cl",  ug/ug-mole.
            7.4.14    Concentration  of  HC1  in  the  flue gas.    Calculate  as
      described below:
                   C = K x m/Vm(std)                                        (3)
            where:  C = Concentration of HC1 or C12, dry basis, mg/dscm,
                    K = 10"3  mg/ug,
                    m = Mass of HC1 or C12 in sample, ug, and
              Vm(std) = Dry gas  volume  measured by  the  dry gas meter,
                        corrected to standard conditions, dscm.
8.0  QUALITY CONTROL
      8.1  Sampling.   See EPA Manual 600/4-77-027b for Method 5 quality control.
      8.2  Analysis.  At the present time, a validated  audit  material  does  not
exist for this method.  Analytical  quality control  procedures are detailed  in
Method 9057.
                                   0050 -  19                       Revision  0
                                                                   January  1995

-------
      8.3  Quality control check sample.  Chloride solutions of reliably known
concentrations  are available  for  purchase  from  the  National  Institute  of
Standards and Technology (SRM 3182).   The  QC check sample should be prepared in
the appropriate absorbing reagent  at  a concentration approximately equal to the
mid range  calibration  standard.   The quality  control check sample  should be
injected in  duplicate  immediately after  the  calibration standards  have  been
injected the first time.   The Cl"  value  obtained for the check sample using the
final  calibration curve should be within 10 percent of the known value for the
check sample.

9.0  METHOD PERFORMANCE

      9.1  The in-stack detection  limit  for the method is approximately 0.02 ug
of HC1  per  liter of stack gas.  The method has a negative bias below 20 ppm HC1
(Reference  6).

      9.2   It  is preferable to include the  cyclone in the  sampling  train to
protect the  filter  from  any moisture present.   There  is  research  in progress
regarding the necessity of  the cyclone  at low moisture  sources  and  the use of
Ascarite II in the drying procedure (Section  7.5.12).

      9.3   The lower detection limit of the  analytical  method  is 0.1 ug of Cl"
per ml of sample solution.   Samples with concentrations which exceed the linear
range of the 1C may be diluted.

      9.4   The  precision  and  bias  for  analysis of HC1  using  this  analytical
protocol have been measured  in combination  with the midget impinger HC1/C12 train
(Method 0051) for sample collection.  The laboratory  relative  standard deviation
is within 6.2 percent and  3.2 percent  at  HC1 concentrations of 3.9 and 15.3 ppm.
respectively.  The method does  not exhibit any bias for HC1 when  sampling at C12
concentrations less than 50 ppm.
                                   0050  -  20                         Revision 0
                                                                  January 1995

-------
10.0  REFERENCES

1.    U. S.  Environmental  Protection Agency, 40 CFR Part 60, Appendix A, Methods
      1-5.

2.    U. S.  Environmental  Protection Agency, "Quality Assurance Handbook for Air
      Pollution Measurement  Systems,  Volume  III,  Stationary Source  Specific
      Methods," Publication No. EPA-600/4-77-027b, August 1977.

3.    Shigehara, R. T., Adjustments in  the EPA Nomography for  Different Pitot
      Tube Coefficients and Dry Molecular Weights, Stack Sampling News,  £:4-ll
      (October 1974).

4.    Steinsberger, S. C.  and J. H. Margeson,  "Laboratory and Field Evaluation
      of  a  Methodology  for Determination of Hydrogen  Chloride  Emissions from
      Municipal  and  Hazardous  Waste   Incinerators,"  U.   S.   Environmental
      Protection  Agency,   Office   of  Research  and  Development,  Report  No.
      600/3/89/064, April, 1989.  Available from NTIS.

5.    State of California, Air  Resources  Board, Method 421,  "Determination of
      Hydrochloric Acid Emissions from Stationary  Sources," March 18,  1987.

6.    Entropy Environmentalists, Inc., "Laboratory  Evaluation of a Sampling and
      Analysis Method  for Hydrogen Chloride Emissions from Stationary Sources:
      Interim  Report,"  EPA Contract No.  68-02-4442,  Research  Triangle  Park,
      North Carolina,  January 22,  1988.
                                  0050  - 21                         Revision 0
                                                                  January 1995

-------
Figure 1.  Isokinetic HC1/C1? sampling  train.
                   0050 -  22
  Revision 0
January 1995

-------
• 2 *


III
b.«A_j  a •«• <^ -3 £ ?


if  l! i  Hi

-------
                           METHOD  0050
  ISOKINETIC  HC1/C1,  EMISSION  SAMPLING  TRAIN
7 1.1 -  7.1,4 Preliminary Procedures:
 Maintain and calibrate all sampling
  equipment, prepare reagents and
        filters (if required}.
7.2.1  Make preliminary measurements
  to establish estimates of isokmetic
      sampling rates and sites.
  7 2 2 - 7 2.3 Select appropriate
  nozzel size to maintain isokinetic
   sampling rate; select apprpnate
   probe liner  and probe lengths.
7 3.1 • 7 3.2 Prepare sampling train,
  add reagents to impingers. Place
  preweighed filter in filter holder if
 particulate matter is to be measured.
   7.3.3 - 7 3.4 Select appropriate
   nozzles and fiber gasket; set-up
     sampling train, check temp.
      settings at ambient temp.
741-7 4.3 Perform leak-check prior
   to sampling run and during run if
     any component is  changed.
                              0050  -   24
                                                    7.5  1 - 7.5.11  Verify filter and probe
                                                    heating temps , maintain an isokmetic
                                                       sampling rate within 10% true
                                                    isokmatic, position nozzel directly  in
                                                       gas stream, initiate sampling.
                                                    7.5.12 At the end of sampling run,
                                                       remove nozzel, turn off pump.
                                                       7 5.13 Recover any liquid m
                                                     cyclone and/or filter by attaching
                                                      an Ascante tube to probe inlet
                                                    and operate tram 1 20 + /- 1 4° C for
                                                                 30 mm.
                                                  7.5.14 - 7 5.15 Perform post-teat leak-
                                                    check, if moisture valve is known,
                                                         calculate % isokmeticity.
                                                                                                              i
                                                    7.6.1  - 7.6.3 Allow probe to cool,
                                                     disassemble umbilical cord, cap,
                                                      and transfer to clean-up area
                                                     764 Collect wash solutions and
                                                        absorbing reagents for field
                                                              blank analysts.
                                                       7 6.6 Remove filter from filter
                                                        holder (if required), place in
                                                        petn dish,  label as container
                                                                  No. 1
                                                    7.6.7  - 7.6.11  Quantitatively recover
                                                    particulate matter from half of probe
                                                    assembly and filter holder.  Combine
                                                     nnses and label as container No. 2.
                                                                                 Revision  0
                                                                                 January  1995

-------
               METHOD  0050  (Cont.)
ISOKINETIC  HC1/C1,  EMISSION  SAMPLING TRAIN
             7.6.12 - 7.6.1 3 Disconnect knockout
             fmpinger and 0.1N H  SO impingers
             (impinger 2 and 3). Record weight or
             volume of solutions, rinse assembly,
              combine solutions and rinses into
              container No. 3. Repeat procedure
            for alkaline impinger solutions, label as
                      container No.4.
               7.6.14 Weigh silica gel from last
                impinger, transfer to original
                        container.
             Analyze impinger solutions for total
                   Cl using Method 9056
                   (Ion Chromatography).
           7.7.1 - 7.7.14 Calculate total particulate
             concentration (7.7.9) and total HCL
              concentration  (7.7.14) in flue as.
                            I
                      0050 -  25
Revision 0
January  1995

-------
                                  METHOD  0051

                MIDGET IMPINGER HC1/C1, EMISSION SAMPLING TRAIN


1.0   SCOPE AND APPLICATION

      1.1   This method describes the collection of hydrogen chloride (HC1,  CAS
Registry Number 7647-01-0) and chlorine  (C12, CAS Registry Number 7782-50-5) in
stack gas emission  samples from hazardous waste  incinerators and municipal waste
combustors.  The collected samples are analyzed using Method 9057.  This method
is designed to  collect  HC1/C12 in their  gaseous forms.   Use  of this method is
limited  to the sampling  of  relatively dry,  particulate-free gas  streams.
Sources, such as those controlled by wet scrubbers,  that emit acid particulate
matter  (e.g.,   HC1  dissolved  in water  droplets)  must  be  sampled using  an
isokinetic HC1/C12  sampling train (see Method 0050).


2.0   SUMMARY OF METHOD

      2.1   An  integrated  gas sample is  extracted  from the  stack  and passes
through a particulate filter, acidified  water,  and finally through an alkaline
solution.  The  filter  serves to remove particulate  matter such  as chloride salts
which could potentially react and form the analyte in the absorbing solutions.
In the acidified water absorbing  solution,  the HC1 gas is solubilized and forms
chloride ions (CV) as follows:

                            HC1  + H20 =  H304 +  CV

The C12  gas present in the emissions  has a very low  solubility in acidified water
and  passes through  to  the  alkaline absorbing  solution  where it  undergoes
hydrolysis to form a proton (H+),  CT, and hypochlorous acid (HC10)  as follows:

                          H20 + C12  = H+  + CT + HC10

The Cl"  ions in the  separate solutions are measured  by  ion chromatography (Method
9057).


3.0   INTERFERENCES

      3.1   Volatile  materials  which produce  chloride  ions  upon  dissolution
during  sampling are  obvious  interferences in the  measurement  of  HC1.   One
interferant for HC1 is diatomic chlorine  (C12) gas which disproportionates to HC1
and hypochlorous acid (HC10)  upon dissolution in water.   C12 gas exhibits a low
solubility in water,  however,  and the use of acidic rather than neutral or basic
solutions for collection of hydrogen chloride gas greatly  reduces the dissolution
of any chlorine present.  Sampling a 400  ppm HC1  gas stream containing 50 ppm C12
with this method does not cause a significant bias.  Sampling  a 200 ppm HC1  gas
                                   0051 - 1                       Revision 0
                                                                  January 1995

-------
stream containing 180 pptn C12 results  in  a positive  bias  of 3.4 percent in the
HC1 measurement.
4.0   APPARATUS AND MATERIALS

      4.1   Sampling  Train.    The  sampling  train  is  shown  in  Figure 1  and
component parts are discussed below.

            4.1.1 Probe.   Borosilicate  glass,   approximately  3/8-in.  (9-mm)
      inside diameter, with a heating  system  to prevent condensation.  When the
      concentration of alkaline particulate matter in the emissions is high,  a
      3/8-in.  (9-mm)  inside  diameter Teflon elbow  should be attached  to the
      inlet of the probe.  A 1-in.  (25-mm) length of Teflon tubing with a 3/8-
      in. (9-mm) inside diameter should be attached at the open end of the elbow
      to permit the opening of the  probe  to be turned away from the gas stream,
      thus reducing the  amount of  particulate  entering  the  train.   When  high
      concentrations of particulate matter are not present, the Teflon elbow is
      unnecessary, and the probe inlet can be perpendicular to the gas stream.
      When  sampling  at  locations  where gas  temperatures  are  greater  than
      approximately 400°F,  such as  wet scrubber  inlets,  glass or  quartz elbows
      must be  used.    In no  case  should  a glass  wool  plug be used  to remove
      particulate matter; use of such  a filtering device  could result in a bias
      in the  data.   Instead,  a Teflon filter  should be  used as  specified in
      Section 5.5.

            4.1.2 Three-way stopcock.  A  borosilicate, three-way glass stopcock
      with a heating system to  prevent condensation.   The  heated stopcock should
      connect directly to  the  outlet  of  the  probe  and  filter assembly and the
      inlet  of  the first  impinger.   The heating system  should  be  capable of
      preventing condensation  up to the  inlet of the first impinger.  Silicone
      grease may be used,  if necessary,  to prevent leakage.

            4.1.3  Impingers.   Five 30-ml  midget impingers with leak-free glass
      connectors.   Silicone  grease   may be used,   if  necessary,   to  prevent
      leakage.  For sampling at high moisture sources or for extended sampling
      times greater than one  hour,  a midget impinger with  a shortened stem  (such
      that the  gas sample does  not  bubble  through  the  collected  condensate)
      should be used  in  front  of the first impinger.

            4.1.4 Mae  West impinger or drying tube.   Mae West design impinger
      (or  drying  tube,  if a moisture determination  is  not to  be  conducted)
      filled  with  silica  gel,  or  equivalent,   to  dry the  gas  sample  and to
      protect the dry  gas  meter and pump.

            4.1.5  Sample line.  Leak-free, with compatible fittings to connect
      the last  impinger  to the needle valve.

            4.1.6  Barometer.   Mercury, aneroid,  or  other barometer capable of
      measuring  atmospheric  pressure  within 2.5 mm  Hg  (0.1  in Hg).   In many


                                   0051  - 2                       Revision 0
                                                                  January  1995

-------
cases, the  barometric reading  may  be obtained  from a  nearby National
Weather Service station,  in  which  case the station  value  (which  is the
absolute barometric pressure) shall be requested and an adjustment for the
elevation differences between the weather station and sampling point shall
be applied at  a rate of  minus  2.5 mm Hg (0.1  in.  Hg) per  30  m (100 ft)
elevation increase or vice versa for elevation decrease.

      4.1.7 Purge pump, purge line,  drying  tube,  needle valve, and rate
meter.  Pump capable of purging  sample  probe at 2 liters/min. with drying
tube, filled with silica  gel or equivalent, to protect  pump,  and  a rate
meter, 0 to 5 liters/min.

      4.1.8 Metering  system.  The following  items  comprise the metering
system which is identical  to that used for EPA Method 6 (see Reference 5).

            4.1.8.1   Valve.   Needle valve, to regulate  sample gas flow
      rate.

            4.1.8.2   Pump.   Leak-free  diaphragm  pump,  or equivalent,  to
      pull gas through train.  Install a small  surge tank between the pump
      and  the rate   meter  to   eliminate  the  pulsation effect  of  the
      diaphragm pump  on the rotameter.

            4.1.8.3   Rate meter.   Rotameter,  or  equivalent,  capable  of
      measuring flow  rate to within 2 percent of selected flow rate of 2
      liters/min.

            4.1.8.4   Volume meter.  Dry gas meter, sufficiently accurate
      to measure  the sample  volume  within 2  percent, calibrated  at the
      selected flow rate  and conditions encountered during sampling, and
      equipped with a temperature gauge (dial  thermometer^or equivalent)
      capable of measuring temperature to within 3°C (5.4°F).

            4.1.8.5   Vacuum gauge.  At least 760 mm Hg (30 in. Hg) gauge
      to be used for  leak check of the sampling train.

      4.1.9 Water Bath:  To minimize loss of absorbing solution

4.2   Sample Recovery.

      4.2.1 Wash bottles.  Polyethylene or glass, 500 ml or larger, two.

      4.2.2 Storage  bottles.  Glass,  with  Teflon-lined  lids,  100  ml,  to
store impinger samples (two per sampling run).
                             0051 - 3                       Revision 0
                                                            January 1995

-------
5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests.  Unless otherwise
indicated, it is intended that  all reagents  shall conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are  available.   Other grades may be used,  provided it is
first ascertained that the  reagent  is  of sufficiently  high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Reagent Water.   All references  to  water  in  the  method  refer to
reagent water  as defined  in  Chapter  One unless  otherwise  specified.   It is
advisable to analyze a blank sample of this reagent prior to sampling, since the
reagent blank value obtained during  the  field sample analysis must be less than
10 percent of the sample values.

      5.3   Sulfuric acid  (0.05 M), H2SO,.   Used  as the HC1  absorbing reagent.
To prepare  100  ml,  slowly  add  0.28  ml of concentrated H,S04 to  about 90 ml of
water while  stirring,  and  adjust  the final volume to IOQ ml  using additional
water.   Shake  well  to mix the  solution.   It is advisable to  analyze a blank
sample of this  reagent prior to sampling, since the reagent blank value obtained
during the  field sample analysis must  be  less  than  10 percent of  the sample
values.

      5.4   Sodium hydroxide (0.1 M),  NaOH.   Used  as the C12 absorbing reagent.
To prepare  100  ml,  dissolve  0.40  g of  solid NaOH  in  about 90  ml  of water and
adjust the final volume to  100 ml using  additional  water.  Shake well  to mix the
solution.   It  is  advisable to  analyze a blank sample  of  this  reagent prior to
sampling, since the reagent blank value obtained during the field sample analysis
must be less than 10 percent of the sample value.

      5.5   Filter.  Teflon mat Pallflex® TX40HI75 or equivalent.   Locate in a
glass, quartz,  or Teflon filter holder with  a Teflon filter support in a filter
box heated to 250°F.

      5.6   Stopcock grease.  Acetone-insoluble,  heat-stable silicone  grease may
be used,  if necessary.

      5.7   Silica gel.  Indicating type, 6- to 16-mesh.  If the silica gel has
been used previously,  dry  at 175°C (350°F)  for 2  hours.   New  silica gel may be
used  as  received.   Alternatively,  other  types  of desiccants  (equivalent or
better) may be  used.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   Sample  collection  is described in  this  method.    The  analytical
procedures  are  described in Method 9057.
                                   0051 - 4                       Revision 0
                                                                  January 1995

-------
      6.2   Samples  should  be  stored   in  clearly  labeled,  tightly  sealed
containers between sample recovery and analysis.  They may be analyzed up to four
weeks after collection.
7.0   PROCEDURE

      7.1   Calibration.   Section 3,5.2  of  EPA's Quality  Assurance Handbook
Volume III (Reference 4) may be used as a guide for these operations.

            7.1.1 Dry Gas Metering System.

                  7.1.1.1  Initial calibration.  Before its initial use in the
            field, first  leak  check the metering  system  (sample line,  drying
            tube, if used, vacuum  gauge, needle valve, pump,  rate meter, and dry
            gas meter)  as follows:   plug the  inlet  end  of  the sampling line,
            pull a vacuum of 250 mm (10 in.)  Hg, plug off  the outlet of the dry
            gas meter, and turn off the pump.  The vacuum should remain stable
            for 30 seconds.   Carefully release the  vacuum  from the system by
            slowly removing  the  plug  from the sample  line  inlet.   Remove the
            sampling line (and drying  tube,  if applicable),  and  connect the dry
            gas metering  system to a appropriately sized wet test meter (e.g.,
            1 liter per revolution).  Make three independent calibration runs,
            using  at  least  five  revolutions  of  the  dry  gas  meter  per  run.
            Calculate the calibration factor,  Y  (wet test  meter calibration
            volume divided by  the dry  gas  meter  volume,  with  both volumes
            adjusted to the same  reference temperature and pressure), for each
            run, and average the results.   If any  Y value  deviates by more than
            2 percent from the  average,  the metering  system  is unacceptable for
            use.   Otherwise, use the average  as  the calibration  factor for
            subsequent test runs.

                  7.1.1.2  Post-test calibration check.  After each field test
            series, conduct  a  calibration check as  in Section 7.1.1.1  above,
            except for the following variations:  (a)  the leak check is not to
            be conducted, (b) three or more revolutions of the dry gas meter may
            be  used,  (c)  only  two  independent  runs  need to be  made.   If the
            calibration factor does no deviate by more than 5 percent from the
            initial calibration factor (determined in Section 7.1.1.1), the dry
            gas meter volumes obtained  during the test series  are acceptable.
            If  the  calibration  factor  deviates  by more  than  5  percent,
            recalibrate  the  metering  system  as  Section  7.1.1.1,  and  for the
            calculations, use the calibration factor  (initial or recalibration)
            that yields the lower gas volume for each  test run.

            7.1.2 Thermometer(s).  Prior to each field test, calibrate against
      mercury-in-glass thermometers at ambient temperature.   If the thermometer
      being  calibrated   reads  within  2°C   (2.6°F)  of  the  mercury-in-glass
      thermometer, it is  acceptable.  If not, adjust the thermometer or use an
      appropriate correction factor.


                                   0051 - 5                       Revision 0
                                                                  January 1995

-------
      7.1.3 Rate meter.  The rate meter should be cleaned and maintained
according to the manufacturer's instructions.

      7.1.4 Barometer.   Prior to each  field test, calibrate  against  a
mercury barometer.   The  field barometer should agree within  0.1  in.  Hg
with the mercury barometer.  If it does not,  the field barometer should be
adjusted.

7.2   Sampling.

      7.2.1 Preparation of collection train.  Prepare the sampling train
as follows:  The first or knockout impinger should have a shortened stem
and be left empty  to  condense moisture  in the gas stream.   The next two
midget impingers should  each  be  filled  with 15 ml of 0.05 M  H2SO,.  The
fourth and fifth impingers  should each be filled with 15  ml of  0.1 M NaOH.
Place  a  fresh  charge of  silica  gel,  or  equivalent,   in  the Mae West
impinger (or the drying tube).  Connect the impingers in series with the
knockout  impinger  first, followed by  the  two  impingers  containing the
acidified reagent  and two impingers containing the alkaline reagent, and
the Mae West impinger containing  the  silica  gel.   If the moisture will be
determined, weigh the impinger assembly  to the nearest + 0.5 g and record
the weight.

      7.2.2 Leak check procedures.   Leak check the  probe  and three-way
stopcock  prior to  inserting  the  probe into  the  stack.    Connect the
stopcock to the outlet of  the  probe,  and  connect  the sample line to the
needle valve.  Plug the probe inlet,  turn on the sample pump, and pull  a
vacuum of at least 250 mm Hg (10  in.  Hg).  Turn  off the needle valve, and
note the  vacuum gauge reading.   The vacuum  should  remain  stable for at
least 30 seconds.  Place  the probe  in the stack at the sampling location,
and adjust the  filter heating system at 250°F and the probe and stopcock
heating systems to  a temperature sufficient to prevent water condensation.
Connect the first impinger  to the stopcock, and connect the sample line to
the last  impinger  and the needle valve.  Upon  completion  of a sampling
run, remove the probe from the stack and  leak check as described  above.
If a leak has occurred,  the sampling run must be voided.  Alternatively,
the portion  of the train  behind  the probe  may be  leak checked between
multiple  runs  at  the same  site  as  follows:  Close  the  stopcock to the
first  impinger, and  turn on the  sample  pump.   Pull  a vacuum of at  least
250 mm Hg (10  in. Hg),  turn  off the needle  valve,  and  note the  vacuum
gauge  reading.  The vacuum  should remain stable for at least 30 seconds.
Release the vacuum on the  impinger  train  by turning the stopcock to the
vent position to permit ambient air to enter.   If  this procedure is  used,
the full train leak check described  above must be  conducted following the
final  run and all  preceding sampling runs voided  if a leak has occurred.

       7.2.3 Purge  procedures.   Immediately prior to sampling,  connect the
purge  line to the stopcock and turn  the  stopcock to  permit the purge pump
to purge  the  probe (see  Figure  1A).   Turn on the purge pump, and  adjust
                             0051  - 6                       Revision 0
                                                            January 1995

-------
      the purge rate to  2  liters/min.   Purge for at least  5  minutes  prior to
      sampling.

            7.2.4 Sample collection.   Turn  on sample pump, pull a slight vacuum
      of approximately 25 mm Hg (1 in. Hg)  on the impinger train,  and turn the
      stopcock to permit stack gas to  be pulled through the impinger train (see
      Figure 1C).  Adjust the sampling rate to 2 liters/min.  as indicated by the
      rate meter, and  maintain this  rate within 10 percent during  the entire
      sampling run.   Take readings of  the dry  gas meter,  the dry  gas  meter
      temperature, rate meter,  and vacuum gauge at least once every five minutes
      during the run.   A sampling  time of one hour is recommended.   However, if
      the expected condensate catch for this sampling  run duration will exceed
      the capacity of  the sampling train,  (1) a larger knockout impinger may be
      used  or  (2) two  sequential half-hour runs  may be  conducted.   At the
      conclusion of the sampling run, remove the train  from the stack, cool, and
      perform a leak check as described in Section 7.2.2.

      7.3   Sample recovery.   Following sampling, disconnect the impinger train
from the remaining sampling equipment  at the  inlet to the knockout impinger and
the outlet to the last impinger.   If performing a moisture determination, wipe
off any moisture on the outside of the train and any excess silicone grease at
the inlet and outlet openings; weigh the train to the  nearest  0.5  g and record
this weight.   Then  disconnect the impingers from each other.   Quantitatively
transfer the contents of the first  three impingers (the  knockout impinger and the
two 0.05 M HpS04 impingers) to a leak-free  storage bottle.  Add the water rinses
of each of tnese impingers  and connecting glassware to the storage bottle.  The
contents  of the  impingers  and connecting  glassware  from  the  second set of
impingers (containing  the 0.1  M NaOH) should be recovered  in  a similar manner if
a Clp analysis  is desired.   The sample bottle should be  sealed,  shaken to mix,
and labeled; the  fluid level   should be marked  so  that  if any  sample  is lost
during transport, a correction proportional to the lost  volume can be applied.
Save portions of  the 0.05  M  H2S04  and 0.1  M  NaOH used as impinger reagents as
reagent blanks.   Take  50 ml  of each  and place  in  separate leak-free storage
bottles.  Label and mark the fluid levels as previously  described.

      7.4   Calculations.  Retain  at least one extra decimal  figure beyond those
contained in the  available data in intermediate calculations, and round off only
the final answer appropriately.

            7.4.1 Nomenclature.

                  Bws    = Water vapor in the gas stream,  proportion by
                          volume.

                  M     = Molecular weight of water, 18.0 g/g-mole
                          (18.0 Ib/lb-mole).

                  Pbar   = Barometric pressure at the exit orifice  of the
                          dry gas meter,  mm Hg (in.  Hg).

                  Pstd   = Standard absolute pressure,  760 mm Hg

                                   0051 - 7                       Revision 0
                                                                  January 1995

-------
            P  ,   = Standard absolute  pressure,  760  mm Hg
                    (29.92 in. Hg).

            R     = Ideal gas constant,  0.06236  mm Hg-m3/°K-g-mole
                    (21.85 in. Hg-ft3/°R-lb-mo1e).

            Tm    = Average dry gas meter  absolute temperature,
                    °K (°R).

            Tstd   = Standard absolute  temperature, 293°K (528°R).

            Vtc   = Total volume  of liquid collected in impingers
                    and  silica gel, ml (equivalent to the
                    difference in weight of the  impinger train
                    before and after sampling,  1 mg  = 1 ml).

            Vm    = Dry  gas volume as  measured  by the dry  gas
                    meter, dcm (dcf).

            Vm(std) = Dry  gas  volume measured by the  dry gas meter,
                    corrected to  standard  conditions, dscm (dscf).

            VH(std) = Volume  of water  vapor  in the gas sample,
                    corrected to  standard  conditions, scm  (scf).

            Y     = Dry  gas meter calibration  factor.

            pa    = Density of water,  0.9982 g/mL (0.002201 Ib/mL).

      7.4.2 Sample volume,  dry  basis,  corrected  to  standard  conditions.
Calculate as described below:
               T        P             UP
               'std      rbar           vm rbar
u      =  V  Y   	      	    =  K Y	
Vstd)    Vm      jm    p                T
where:
      = 0.3858°K/mm Hg for metric units.
      = 17.64°R/in. Hg for English units.
      7.4.3 Volume of water  vapor.
            v      -V         --.- =  KV               (2)
            Vstd)   Vlc                     "Vic             ^1
                         Mw        Pstd
                              0051  -  8                       Revision 0
                                                             January 1995

-------
where:

K2    = 0.0013333 m3/ml_ for metric units.
      = 0.04707 ft3/mL for English units.

      7.4.4 Moisture content.


            Bus	*«!»----                                 (3)
                  V      + V
                  Ym(std)  T 'w(std)


      7.4.4  Analytical  calculation of total ug HC1 per sample.  Calculate
as described below:

              mHcl =  (S-B)  x  Vs x  36.46/35.45                        (1)

      where:  mHcl =  Mass  of  HC1  in  sample,  ug,

                 S = Analysis of  sample, ug CV/mL,

                 B = Analysis of  reagent blank, ug CV/mL,

                Vs = Volume of filtered and diluted sample, ml,

             36.46 = Molecular weight of HC1, ug/ug-mole,  and

             35.45 = Atomic weight of Cl",  ug/ug-mole.


      7.4.5  Analytical  calculation of total ug C12 per sample.  Calculate
as described below:

              MC12 =  (S-B)  x  Vs x  70.91/35.45                        (2)

      where:  MC12 =  Mass  of  C12 in sample,  ug,

             70.91 = Molecular weight of C12,  ug/ug-mole, and

             35.45 = Atomic weight of Cl",  ug/ug-mole.

                 S = Analysis of  sample, ug CV/mL,

                 B = Analysis of  reagent blank, ug CV/mL,
                             0051 - 9                       Revision 0
                                                            January 1995

-------
      below:
7.4.6  Concentration of HC1 in the flue gas.  Calculate as described


       c    = K x m/Vm(std)                                  (3)

where:  C   = Concentration of HC1 or C12,  dry basis,  mg/dscm,

        K   = 10"3 mg/ug,

        m   = Mass of HC1 or C12 in sample,  ug,  and

  Vm(std)     = Dry gas volume measured by the dry gas meter,
            corrected to standard conditions,  dscm.
8.0   QUALITY CONTROL

      8.1   At the present time, a validated audit material does not exist for
this method.  Analytical  quality control procedures are detailed in Method 9057.

      8.2  Quality control check sample.  Chloride solutions of reliably known
concentrations are available for purchase from  the National Bureau of Standards
(SRM 3182).  The  QC check sample should be prepared in the appropriate absorbing
reagent  at  a concentration  approximately  equal  to  the mid  range calibration
standard.   The  quality  control  check  sample should  be  injected  in duplicate
immediately after the calibration standards have been  injected the first time.
The Cl" value obtained for the  check  sample using the final  calibration curve
should be within 10 percent of the known value for the check sample.
9.0   METHOD PERFORMANCE

      9.1   The in-stack detection limit  for the method is approximately 0.08 ug
of HC1 per liter of stack gas for a 1-hour sample.

      9.2   The precision and bias  for  measurement of HC1 using this  sampling
protocol  combined  with  the  analytical  protocol  for Method  0050  have been
determined.  The laboratory relative standard deviation is within 6.2 percent and
3.2 percent at HC1  concentrations  of 3.9 and 15.3 ppm,  respectively.   The method
does not exhibit
50 ppm.
     any bias for HC1  when sampling  at  C12 concentrations less than
10.0  REFERENCES
      Steinsberger, S.C. and J.H. Margeson,  "Laboratory and  Field Evaluation of
      a  Methodology  for  Determination  of Hydrogen  Chloride  Emissions  from
      Municipal and Hazardous Waste Incinerators," U.S.  Environmental Protection
      Agency,  Office of  Research and  Development,  Report  No.  600/3/89/064,
      April, 1989.  Available from NTIS.
                                   0051  - 10
                                                      Revision 0
                                                      January 1995

-------
2.    State of California, Air  Resources  Board,  Method 421,  "Determination of
      Hydrochloric Acid Emissions from Stationary Sources," March 18, 1987.

3.    Entropy Environmentalists, Inc.,  "Laboratory  Evaluation of a Sampling and
      Analysis Method for Hydrogen Chloride Emissions from Stationary Sources:
      Interim Report,"  EPA Contract No.  68-02-4442, Research Triangle  Park,
      North Carolina, January 22, 1988.

4.    U.S. Environmental  Protection  Agency,  "Quality  Assurance Handbook for Air
      Pollution Measurement  Systems,  Volume  III,  Stationary Source Specific
      Methods," Publication No.  EPA-600/4-77-027b,  August 1977.

5.    U.S. Environmental  Protection Agency, 40 CFR Part 60, Appendix A, Method
      6.
                                  0051  -  11                       Revision 0
                                                                  January 1995

-------
                                                       i
                                                        i
                                             WIW
           HCtOj nm0hif Ma
0051 - 12
Revision  0
January  1995

-------
                             METHOD  0051
MIDGET  IMPINGER  HC1/C1:  EMISSION  SAMPLING  TRAIN
                       7.1.1.1 Leak-check dry gas metering
                      system, calculate calibration factor Y.
                          7.1.1.2 Poat teat calibration
                            after each teat aeries.
                       7.1.2 - 7.1.4 Calibrate termometers,
                           rate meter and barometer.
                      7.2.1 Set up sample collection train,
                      connect impmgers in series, and add
                         reagents. If moisture la to be
                     determined, weigh Mae Weat impinger
                                with silica gel.
                     7.2.2 Perform leak-check on assembled
                                sampling train.
                      7.2 3 Immediately prior to sampling,
                      purge probe for 5 mm. at 2 liter/mm.
                         7.2.4 Perform aample collection,
                       adjust aampling rate to 2 litera/min.
                       for 1 hour. Take readings for dry ges
                       meter, tempsratura, rate meter, and
                           vaccum once every 5 mm.
                      7.3 Stop pump, disconnect tmpmger
                      sampling tram, quantitatively transfer
                     contents of impmgers and water rinses
                       to labeled  containers. Retain 50 ml
                          ah quota of 0.1N H 2SO 4and
                              NaOH for blanks.
                      7.4 Calculate ssmpls volume (7.4.2),
                      volume of wear vapor (7.4.3), total ug
                      HCL per sample (7.4.4), ug CL 2 per
                         sample {7.4.5}, and total HCL
                             concentration (7.4.6).
                                0051  -  13
   Revision  0
January  1995

-------
                                  METHOD 0100

            SAMPLING FOR FORMALDEHYDE AND OTHER CARBONYL COMPOUNDS
                                 IN INDOOR AIR
1.0   SCOPE AND APPLICATION

      1.1   This method provides procedures for the sampling of various carbonyl
compounds in indoor air by derivatization with 2,4-dinitrophenylhydrazine (DNPH)
in a silica gel cartridge.   The method  may  be  used in conjunction with Method
8315.  The following compounds may be sampled by this method:
            Compound Name                              CAS No."


            Acetaldehyde                                75-07-0
            Acetone                                     67-64-1
            Acrolein                                   107-02-8
            Benzaldehyde                               100-52-7
            Butyraldehyde                              123-72-8
            Crotonaldehyde                             123-73-9
            2,5-Dimethylbenzaldehyde                  5779-94-2
            Formaldehyde                                50-00-0
            Hexanal                                     66-25-1
            Isovaleraldehyde                           590-86-3
            Propionaldehyde                            123-38-6
            m-Tolualdehyde                             620-23-5
            o-Tolualdehyde                             529-20-4
            p-Tolualdehyde                             104-87-0
            Valeraldehyde                              110-62-3


      8  Chemical  Abstract Services Registry Number

      1.2   This method is restricted to use by, or under the close supervision
of, trained analytical  personnel  experienced in  sampling  organic compounds in
air.  Each analyst must demonstrate the ability to generate acceptable results
with this method.
2.0   SUMMARY OF METHOD

      2.1   A known volume of indoor air is drawn through  a prepacked silica gel
cartridge  coated  with  acidified  2,4-dinitrophenylhydrazine  (DNPH),  at  a
predetermined sampling rate for an appropriate period of time.  After sampling,
the sample cartridges  are capped and placed in  borosilicate  glass  tubes with
polypropylene caps and placed in cold  storage until analysis.  The compounds of


                                   0100 - 1                         Revision 0
                                                                  January 1995

-------
interest may then be eluted from the cartridge with acetonitrile from a plastic
syringe reservoir into a graduated test tube or a volumetric flask.
The eluate  is then  topped to  known  volume  and refrigerated  until  analysis.
Analysis may be done  High Performance Liquid Chromatography  (HPLC), Method 8315,
with an ultraviolet  (UV/Vis) detector at 360 nm.


3.0   INTERFERENCES

      3.1   Solvents, reagents,  glassware and other sample processing may yield
discrete artifacts and/or  elevated  baselines causing  misinterpretation of the
chromatograms.   All  of these materials  must  be demonstrated to  be  free from
interferences, under the conditions of analysis, by analyzing method blanks.

            3.1.1   Glassware and  plasticware   must  be  scrupulously  cleaned.
      Clean  all  glassware and  plasticware as   soon as  possible  after  use  by
      rinsing with  the last solvent used  in it.   This  should  be followed  by
      detergent washing with hot water and  rinsing with tap water, organic-free
      reagent water,  and aldehyde-free acetonitrile.  After cleaning, glassware
      and plasticware  should  be stored in a clean environment  to prevent any
      accumulation of dust or other contaminants.

            3.1.2  High purity reagents and solvents should be used to minimize
      interference problems.  Purification of solvents by distillation in all-
      glass systems may be necessary.

            3.1.3  Polyethylene gloves should be worn when handling the silica
      gel cartridges to reduce  the possibility of contamination.

      3.2   Contamination  of the  DNPH  reagent is  a frequently encountered
problem.  Formaldehyde, acetone, and 2,4-dinitroanaline (a decomposition product
of DNPH) may  be  significant  analytical  impurities  in  the DNPH reagent at high
concentrations.  The DNPH  must be purified by multiple  recrystallizations  in UV-
grade  acetonitrile.    Recrystallization  is accomplished, at  40-60°C,  by slow
evaporation of the solvent to maximize crystal  size.  The  purified  DNPH crystals
are stored under UV-grade  acetonitrile until use.  Impurity levels of carbonyl
compounds in  the DNPH  are  determined  prior to  the  analysis of the samples and
should  be  less  than 0.025 /jg/mL.   Refer to Sec. 5.9  for  a recrystallization
procedure.

      3.3   Ozone Interferences - Ozone at high  concentration  has  been shown to
interfere negatively by reacting with both DNPH  ar.d  its hydrazone derivatives in
the cartridge (Ref.  6).

            3.3.1  The extent of interference depends on the temporal variations
      of  both the  ozone  and  the  carbonyl  compounds during sampling.   The
      presence  of ozone  in  the  sample  stream  is readily inferred  from the
      appearance of new compounds with retention times shorter than that of the
      hydrazone of formaldehyde.  Figure 1 shows chromatograms of samples of a
      formaldehyde-spiked  air stream with  and without ozone.


                                   0100  -  2                         Revision 0
                                                                  January 1995

-------
            3.3.2   The  most direct solution  to  the ozone  interference  is  to
      remove the ozone  before  the sample stream reaches the  cartridge.   This
      process entails constructing an ozone denuder or scrubber and placing it
      on the front of the cartridge.  The denuder is constructed out of 1 m of
      0.64 cm OD copper  tubing,  which  is  filled with a saturated solution of KI
      water, allowed  to stand  for  approximately  5  minutes, and dried  with  a
      stream of clean air  or nitrogen for about 1 hour.   The  capacity  of the
      ozone denuder as  described  is  about  10,000  ppb/hour  of ozone.    Test
      aldehydes that were dynamically spiked into an ambient sample air stream
      passed through the denuder with virtually no losses.

      3.4   Samples may  be contaminated during shipment or  storage by diffusion
of volatile organics through the sample bottle septum seal.  Field reagent blanks
must be analyzed to determine when sampling and storage procedures have caused
the contamination.

      3.5   Matrix interferences may be caused by contaminants acquired by the
sampling process.  The extent of matrix interferences will  vary considerably from
source to source, depending  upon  the  nature  and  diversity  of  the  matrix being
sampled.  If significant interferences occur  due to organic compounds that have
the same retention time, altering the separation conditions  by using alternative
HPLC columns or mobile phase conditions may resolve the problem.


4.0   APPARATUS AND MATERIALS

      4.1   Sampling Equipment

            4.1.1   Sampling  System  -   capable  of  accurately and  precisely
      sampling 0.10 to  1.50 L/min of indoor  air.   The  procedures  given here
      assume use of a  dry meter-equipped  sampling system operated at flow rates
      of at least 0.5 L/min.

NOTE: A normal  pressure  drop through the sample cartridge approaches 19 kPa at
      a sampling rate of 1.5 L/min.

            4.1.2  Thermometer and  Barometer  - to record indoor conditions  at
      the time of sampling.

            4.1.3  Stopwatch -  to time sampling.

            4.1.4  Rotameters - to  allow observation of  the flow  rate without
      interruption of the sampling process.

            4.1.5  Mass  Flowmeters and Mass Flow Controllers -  for metering and
      setting the  air flow  rate  through the sample  cartridge (0.50 to  1.20
      L/min).   These are necessary because cartridges have  a high pressure drop
      and,  at  maximum  flow rates,  the  cartridge  behaves like  a  "critical
      orifice"  and  can display a flow rate drop over an extended sampling period
      (generally less than 5% over a 24  hour period).


                                   0100  - 3                         Revision  0
                                                                  January 1995

-------
            4.1.6   Fittings  and Plugs  (Luer-Lok  or equivalent) -  to  connect
      cartridges to the sampling system and to cap prepared cartridges.

            4.1.7  Heated Probe - necessary when the temperature of sampled air
      is  below 60"C,  to  insure  effective  collection  of  formaldehyde  as  a
      hydrazone.

            4.1.8  Silica Gel  Cartridges - chromatographic grade, 2 cm x 1.5 cm
      ID, with  Luer-Lok  type  fittings on each end, for manual  application of
      acidified DNPH  coating  (Sep-PAK from Waters Associates  or  equivalent).
      Commercially  pre-packaged  pre-coated  cartridges  are  also  available
      (Thermosorb/F cartridges from Thermedics Inc. or equivalent).

      4.2   Glassware

            4.2.1  Volumetric Flasks - various sizes,  5 to 2000 mL.

            4.2.2  Pipets - various sizes,  1 to 50 mL

            4.2.3  Sample Vials

            4.2.4    Borosilicate  glass  culture  tubes  (20 x  125  mm)  with
      polypropylene screw caps - for transporting coated cartridges.

      4.3   Liquid Syringes (polypropylene are adequate)  - 10 mL,  used to prepare
DNPH-coated cartridges.

      4.4   Syringe Rack - made of an aluminum plate with adjustable legs on all
four corners.   Circular holes of a diameter slightly larger than the diameter of
the 10 mL syringes  are drilled through  the  plate  to allow batch processing of
cartridges for cleaning,  coating, and  sample elution.  A 0.16 x 36 x 53 cm plate
with 45 holes in a 5x9 matrix  is recommended.  See Figure 2.

      4.5   Cartridge  Drying  Manifold - has  multiple standard male  fittings
(Luer-Lok or equivalent).  See Figure 2.

      4.6   Repetitive Dispensing  Pipets  -  positive displacement,  0 to  10 mL
range, with 1 L reagent bottles (Lab-Industries or equivalent).

      4.7   Polyethylene Gloves - used to handle silica gel  cartridges.

      4.8   Sample Vial Holder - Friction-top metal can (e.g., 4  L paint can) or
a styrofoam box lined  with either  polyethylene  air bubble  padding  or granular
charcoal to cushion the samples.

      4.9   Soap Bubble Flowmeter or Calibrated Wet Test Meter - for calibrating
the sampling flow rate.

      4.10  Melting Point Apparatus (optional)
                                   0100 - 4                         Revision 0
                                                                  January 1995

-------
5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests.  Unless otherwise
indicated, it is intended that  all reagents  shall conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are  available.   Other  grades  may be used,  provided it is
first ascertained that the reagent is  of sufficiently  high purity to permit its
use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Nitrogen gas, N2 -  high  purity  grade.

      5.4   Acetonitrile, CH3CN - UV  grade.

      5.5   Formaldehyde, CH20  -  ACS certified or assayed 36.5% solution (W/W).

      5.6   Aldehydes and Ketones - analytical grade,  used  for preparation of
DNPH derivative standards of target analytes  other than formaldehyde.  See list
on page 1 for possible target analytes.

      5.7   Perchloric Acid, HC104  -  analytical grade.

      5.8   Hydrochloric Acid,  HC1  -  analytical  grade.

      5.9   2,4-Dinitrophenylhydrazine (DNPH),  C6H6N404 - recrystallize at least
twice with UV grade acetonitrile using the  following procedure:

NOTE: This procedure  should be  performed  under  a  properly  ventilated hood.
      Inhalation of acetonitrile  can result in nose and throat  irritation (brief
      exposure  at  500 ppm)  or more  serious  effects  at higher concentration
      and/or longer exposures.

            5.9.1  Prepare a saturated solution of DNPH by boiling excess DNPH
      in 200 ml of acetonitrile for approximately  1  hour.

            5.9.2   After 1  hour,  remove  and  transfer the  supernatant  to  a
      covered beaker on  a hot  plate  and allow gradual cooling  to  40  to 60°C.
      Maintain this temperature  range until  95% of  the  solvent has evaporated
      leaving crystals.

            5.9.3  Decant the solution to waste and rinse the remaining crystals
      twice with three times their apparent volume of acetonitrile.

            5.9.4   Transfer the crystals  to a clean beaker,  add 200 ml  of
      acetonitrile, heat to boiling,  and again let the crystals grow slowly at
      40 to 60°C until 95%  of  the solvent  has evaporated.   Repeat the rinsing
      process as in Sec. 5.9.3.
                                   0100 - 5                         Revision 0
                                                                  January 1995

-------
            5.9.5  Take  an  aliquot  of the second rinse, dilute  10  times with
      acetonitrile, acidify with  1  ml of 3.8 M perchloric acid  per  100  ml of
      DNPH solution,  and analyze by HPLC Method 8315).   An  acceptable impurity
      level is  less  than 0.025  mg/L of  formaldehyde  in  recrystallized DNPH
      reagent or below the sensitivity (ppb, v/v) level  indicated  in Table 1 for
      the anticipated sample volume.

            5.9.6  If the impurity concentration is not satisfactory, pipet off
      the solution to waste, repeat the recrystallization as  in Sec.  5.9.4 but
      rinse with  two 25 ml  portions of acetonitrile.   Prep and analyze  the
      second rinse as in Sec. 5.9.5.

            5.9.7  When  the  impurity concentration j_s  satisfactory,  place the
      crystals  in  an   all-glass   reagent   bottle,  add   another   25  mL  of
      acetonitrile,  stopper,  and shake  the bottle.   Use  clean pipets when
      removing the saturated DNPH stock  solution  to  reduce the possibility of
      contamination of  the  solution.  Maintain only a minimum  volume of the
      saturated solution adequate for day to  day operation to minimize waste of
      the purified reagent.

      5.10  Refer  to  the determinative  method (Method  8315)  for  procedures
regarding the  preparation of DNPH derivatives, standards of  the derivatives, and
calibration standards for HPLC analysis.  All standard solutions  should be stored
at about 4°C in a glass vial with a Teflon®-!ined cap,  with minimum headspace,
and in the dark.   They should be stable for about 6 weeks.  All standards should
be checked frequently for signs of degradation or evaporation,  especially just
prior to preparing calibration standards from them.

      5.11  Preparation of DNPH-Coated Sep-PAK  Cartridges (if pre-packaged pre-
coated cartridges, as in Sec. 4.1.8, are not used)

NOTE: This procedure  must be performed in an atmosphere  with a very low aldehyde
      background.  The atmosphere above the acidified  solution should preferably
      be  filtered through  a  DNPH-coated  silica  gel  cartridge to  minimize
      contamination from laboratory  air.  All glassware and plasticware must be
      scrupulously cleaned  and  rinsed with deionized water  and  aldehyde free
      acetonitrile.  Contact of reagents  with laboratory air must be minimized.
      Polyethylene gloves must be worn when handling the cartridges.

            5.11.1  DNPH Coating Solution

                  5.11.1.1  Pipet 30 ml  of  saturated  DNPH stock solution  into a
            1000 mL volumetric flask, add 500 mL acetonitrile, and acidify with
            1.0 mL of concentrated HC1.

                  5.11.1.2     Shake  solution   and  dilute  to  volume   with
            acetonitrile.  Stopper  the  flask,  invert,  and  shake  several  times
            until the solution is homogeneous.  Transfer the acidified solution
            to a  reagent bottle equipped with  a  0  to 10  mL range  repetitive
            pipet dispenser.  Prime the dispenser and slowly dispense 10 to 20
            mL to waste.

                                   0100 - 6                         Revision 0
                                                                  January 1995

-------
      5.11.1.3  Dispense an aliquot solution to a sample vial,  and
check the impurity level of the acidified  solution by HPLC according
to Sec. 7.2.

      5.11.1.4  The  impurity concentration should be less than 0.025
jug/mL formaldehyde, similar to that in the DNPH stock solution.

5.11.2  Coating of Sep-PAK Cartridges

      5.11.2.1  Open the Sep-PAK package, connect the short end to
a 10 ml syringe and  place  it  in the syringe rack.  The syringe rack
used for coating and drying the sample cartridges is illustrated in
Figures 2(a) and 2(b).

      5.11.2.2  Using  a positive displacement,  repetitive pipet,  add
10 ml of acetonitrile to each of the syringes.

      5.11.2.3  Let the liquid drain to  waste by gravity.   Remove
any air  bubbles that  may be  trapped  between the syringe  and  the
silica cartridge by displacing  them with the  acetonitrile  in  the
syringe.

      5.11.2.4   Once  the  effluent  flow  at  the outlet  of  the
cartridge has stopped, dispense 7 ml of the acidified DNPH coating
reagent  into each  of  the  syringes  using  the repetitive  pipet
dispenser.

      5.11.2.5  Let  the coating reagent drain by  gravity through the
cartridge until flow at the other end of the cartridge stops.

      5.11.2.6  Wipe the excess  liquid  at the  outlet of each of the
cartridges with clean tissue paper.

      5.11.2.7  Assemble a drying manifold as  shown in Figure 2(b).
This contains a previously prepared, DNPH-coated, cartridge at each
of the exit  ports (e.g., these scrubber or "guard cartridges" can be
prepared by drying a few of the newly coated cartridges as per  the
following sections,  and  "sacrificing" these few to assure the purity
of the  rest).   The "guard cartridges"  serve to remove  traces of
formaldehyde that may be present in the nitrogen gas  supply.

      5.11.2.8  Insert  cartridge connectors  (flared  at both  ends,
0.64 cm OD x 2.5 cm Teflon® FEP tubing with  ID  slightly smaller than
the OD  of the cartridge  port) onto  the  long end of  the scrubber
cartridges.

      5.11.2.9  Remove the cartridges from the syringes and connect
the short ends  of the cartridges to the  open  end  of  the cartridge
connectors already attached to the scrubber cartridges.
                       0100 - 7                         Revision 0
                                                      January 1995

-------
                  5.11.2.10   Pass  nitrogen  through each of the  cartridges  at
            about 300 to 400 mL/min.

                  5.11.2.11  Rinse the exterior surfaces and outlet end of the
            cartridges with acetonitrile using a Pasteur pipet.

                  5.11.2.12  After 15 minutes, stop the flow of nitrogen,  wipe
            the cartridge  exterior free of rinse  acetonitrile and  remove the
            dried cartridges.

                  5.11.2.13   Plug  both  ends of  the  coated  cartridge  with
            standard polypropylene Luer-Lok  male plugs and place  the plugged
            cartridge in a  borosilicate  glass culture tube with  polypropylene
            screw caps.

                  5.11.2.14  Put a serial  number  and a  lot number label on each
            of the individual cartridge glass storage containers and refrigerate
            the  prepared  lot  until  use.    Cartridges will  maintain  their
            integrity for up to 90 days stored in refrigerated,  capped culture
            tubes.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Assemble the sampling system,  and ensure that the pump is capable of
constant flow rate throughout the sampling period.  The coated cartridges can be
used as direct probes and traps  for sampling  indoor air when the temperature is
above 60°F.

            6.1.1   If the  temperature  in below  60°F,  use the  heated probe,
      mentioned in Sec.  4.1.7, to warm the air entering the sampling equipment.

            6.1.2  If necessary, add an ozone denuder (see Sec.  3.3).

      6.2   Before sample collection, check the system  for leaks.  Plug  the input
(short end) of the cartridge  so no flow is  indicated  at the output end of the
pump.  The mass flowmeter should not  indicate  any  air  flow through the sampling
apparatus.

      6.3   Install the  entire assembly (including a "dummy" sampling cartridge)
and check the flow rate at a value near the desired rate.  In  general, flow rates
of 500-1200 mL/min should be employed.  The total moles of carbonyl in the volume
of air sampled should not exceed that  of  the DNPH (2 mg  or 0.01 mmole/cartridge).
In general, a safe estimate of the sample size  should be approximately 75% of the
DNPH  loading  of  the cartridge  (approximately  200 /zg  as  HCHO).   Generally,
calibration is accomplished using a soap bubble flowmeter or calibrated wet test
meter connected to the flow exit, assuming the system is sealed.

NOTE: ASTM Method D3686 describes  an  appropriate calibration  scheme  that does
      not require a  sealed flow system downstream of the pump.


                                   0100 - 8                         Revision  0
                                                                  January 1995

-------
      6.4   Ideally, a dry gas meter  is  included  in  the  system to measure and
record total  flow.   If  a  dry gas meter  is  not available, the  operator must
measure  and  record the  sampling  flow rate  at  the  beginning  and end  of the
sampling period to determine  sample  volume.   If  the sampling period exceeds two
hours, the  flow  rate  should be  measured at intermediate points during the
sampling period.   Include a  rotameter to allow observation of  the  flow rate
without interruption of the sampling process.

      6.5   Before sampling,  remove the glass culture tube from the friction-top
metal can or styrofoam box.   Let the cartridge warm to room temperature in the
glass tube before connecting  it to the sample train.

      6.6   Using polyethylene gloves, remove the coated cartridge from the glass
tube and  connect it to the sampling system with a Luer adapter fitting.  Seal the
glass tube for later use, and connect the cartridge to the  sampling train  so that
the short end becomes the sample inlet.

      6.7   Turn the sampler  on, record the start time,  and adjust the flow to
the desired  rate.  A typical flow rate  through one cartridge is  1.0 L/min  and 0.8
L/min for two cartridges in tandem.

      6.8   Operate the sampler for  the desired  period, with periodic recording
of the sampling variables such as sample flow rate, pressure,  and temperature.

      6.9   At the end  of the  sampling period, stop the flow and record the stop
time.  If  a dry gas meter or equivalent is not  used,  the flow  rate  must be
checked just  before  stopping the flow.   The average sample flow  rate  may be
calculated using the equation in Sec.  9.1.1.  If the  flow  rate at the beginning
and end of  the sampling  period differ by more  than  15%,  the  sample  should be
marked as suspect.

      6.10  Immediately after sampling, remove the cartridge (using polyethylene
gloves) from the sampling system,  cap  with Luer  end plugs,  and place it back in
the original  labeled glass culture  tube.   Cap  the culture  tube,  seal  it with
Teflon® tape,  label  the tube,  and  place it in a  friction-top can containing 2-5
cm of granular charcoal  or styrofoam box with appropriate  padding.  Refrigerate
the culture tubes until  analysis.   The refrigeration period prior to analysis
should not exceed 30 days.

NOTE: If samples  are to  be shipped  to a  central  laboratory for  analysis, the
      duration  of  the  non-refrigerated  period  should be kept to a  minimum,
      preferably less than two days.

      6.11   Use the equations found in Sees. 9.1.2  and 9.1.3  to  calculate the
total volume  of air sampled   and  the  total  volume of air sampled at standard
conditions.
                                   0100 - 9                         Revision 0
                                                                  January 1995

-------
7.0   SAMPLE RECOVERY

      7.1   The samples  are  returned  to the laboratory  in  a  friction-top can
containing 2 to 5  cm of granular charcoal and  stored  in a  refrigerator until
analysis.   Alternatively,  the  samples may  also  be  stored  alone  in  their
individual glass containers.   The time  between sampling and analysis should not
exceed 30 days.

      7.2   Refer  to the  determinative method  (Method  8315)  for  procedures
regarding  desorption of the  sample   from  the  cartridge  and  HPLC  analysis
preparation.


8.0   CALIBRATIONS

      8.1   Refer to Sec. 6.0 for requirements regarding  the calibration of the
sampling system flow rate and equipment for the determination of total flow.

      8.2   Refer  to   the   determinative   method  for  procedures  regarding
calibration of the HPLC analysis system.

      8.3   Barometer -  Adjust the barometer  initially  and  before  each test
series to agree within ±2.5 mm Hg (±0.1 in Hg) of the mercury barometer or the
corrected barometric pressure value reported by a nearby National  Weather Service
Station  (same  altitude  above sea level).   Note  that  adjustment for elevation
differences between the weather station and the sampling point  is applied at a
rate of minus 2.5 mm Hg  (0.1 in Hg) per 30 m (100 ft) elevation increase.

      8.4   Thermometers

            8.4.1  If a  mercury-in-glass reference thermometer is to  be used, it
      must conform to ATSM E-l 63C or 63F specifications.

            8.4.2  If a  thermocouple is  to be used, it must be  calibrated  in the
      laboratory   according   to  the    manufacturer's  specifications.     The
      calibration should be done both with  and without  the use of any extension
      leads.
9.0   CALCULATIONS

      9.1  Calculation of the total  volume of air sampled at standard conditions.

            9.1.1  If a dry gas meter or equivalent total flow indicator is not
      used, the average sample flow rate, FRave in mL/minute, may be calculated
      using the following equation:

                                rr\i •+• ri\2 +  ...  + rKpj
                        r K.,..-  = 	
                                   0100 -  10                        Revision 0
                                                                  January 1995

-------
            where:

                 FR,, FR2, ..., FRN  =  Flow  rates  determined at the beginning,
                                       end,  and  intermediate points during
                                       sampling
                                   N  =  Number  of flow rates averaged

            9.1.2  The total  volume  of air  sampled  at the measured temperature
      and pressure,  VTot  in liters (L),  may be  calculated  using  the following
      equation:

                                 (Time2  - Time,)  (FRave)
                                         1,000

            where:
                        Time2 = Stop time  (min)
                        Time! = Start time  (min)
              (Time2 - Time.,) = Total sampling time  (min)
                        FRave  = Average  flow rate  (mL/min)

            9.1.3   The  total  volume  of  air  sampled  converted  to  standard
      conditions, VTotStd in  liter  (L)  at  25°C and 101.3  kPa,  may be calculated
      using the following equation:

                                         Pave           298°C
                      V       = V   "           "
                      * TotStd     ' Tot
                                       101.3  kPa    (273°C + Tave)

            where:
                   VTot  = Total  sample  volume  (L)  at  measured temperature and
                          pressure
                   Pa*e  = Average indoor  pressure  (kPa)
                   Lve  = Average indoor  temperature  (°C)


10.0  DETERMINATION OF VOLUME TO BE SAMPLED

      10.1  Refer to Table  1 for information regarding method "sensitivity" at
various sampling volumes.


11.0  QUALITY CONTROL

      11.1  Refer to Chapter One for quality control  procedures.

      11.2  Method  Blanks  - A  method  blank must  be  prepared for each  set of
analytical  operations,  to  evaluate  contamination  and   artifacts  that  can  be
derived from glassware, reagents, and  sample handling in the  laboratory.


                                   0100 -  11                         Revision 0
                                                                   January 1995

-------
      11.3  Field Blanks  -  Field  blanks  must  be  submitted  with  the samples at
each sampling site or 10% of the field samples, whichever is larger, should be
shipped and analyzed with each  group of samples.  The  field  blank is treated
identically to the samples  except that no  air is drawn  through  the cartridge.
It is desirable to analyze blank cartridges retained in  the laboratory (method
blanks)  as  well,  to  distinguish  between  possible  field  and  laboratory
contamination.

      11.4  Blank and  Matrix Spikes  -  A  procedure  for spiking  air  sampling
cartridges  is not yet established  for  this  sampling  technique.    Refer  to
Attachment A  for  information regarding  possible  techniques  for accomplishing
sample spiking.  Proper QC procedures require that a blank spike and matrix spike
be processed  for each  batch  of  10 samples  or  less.   As  the MDL  becomes better
established for this method, the  representative spike concentration should be set
at 10 times the MDL,  for that matrix, to account for interferences.


12.0  METHOD  PERFORMANCE

      12.1  The  method  detection  limit  (MDL)   is  defined  as  the  minimum
concentration of the test compound  that  can  be  measured and  reported with 99
percent confidence as being greater than zero.  The MDL  actually achieved in a
given analysis will vary, as it is dependent on instrument sensitivity and matrix
effects.   The MDLs  for  the target  analytes  in  the  method have  not  yet  been
established.

      12.2  Table 1  illustrates  the  sensitivity for the  target  analytes  of
interest found in ambient air that  have been  identified using  two Zorbax CDS
columns in series.
13.0  REFERENCES

1.    Winberry,  Jr.,  W.T.,  Murphy,  N.T.,  and  Riggin,  R.M.,  Method  TO-11,
      Compendium of Methods For the Determination of Toxic Organic Compounds in
      Ambient  Air,  EPA-600/6-89-017,  U.S.  Environmental  Protection  Agency,
      Research Triangle Park, NC, June 1988.

2.    Tejada,  S.B.,  "Standard  Operating   Procedure  for  DNPH-coated  Silica
      Cartridges For Sampling  Carbonyl Compounds  in  Air and  Analysis by High-
      performance   Liquid  Chromatography,"  Unpublished,  U.S.  Environmental
      Protection Agency, Research Triangle Park, NC,  March 1986.

3.    Tejada,  S.B., "Evaluation  of Silica Gel Cartridges Coated  in  situ with
      Acidified 2,4-Dinitrophenylhydrazine for Sampling Aldehydes and  Ketones in
      Air,"  Intern. J. Environ. Anal. Chem., Vol. 26:167-185, 1986.

4.    Quality Assurance Handbook for Air Pollution Measurement Systems, Volume
      II - Ambient  Air Specific Methods,  EPA-600/4-77-027A,  U.S. Environmental
      Protection Agency, Research Triangle Park, NC,  July 1979.


                                   0100 -  12                         Revision 0
                                                                  January 1995

-------
5.    Riggin, R.M., Technical Assistance Document for Sampling and Analysis of
      Toxic  Organic   Compounds  in   Ambient  Air,   EPA-600/4-83-027,   U.S.
      Environmental Protection Agency, Research Triangle Park, NC, June, 1983.

6.    Arnts, R.R.  and  Tejada,  S.B.,  "2,4-Dinitrophenylhydrazine-Coated Silica
      Gel Cartridge Method for Determination  of  Formaldehyde  in Air", Env. Sci.
      and Tech. 23, 1428-1430 (1989).
                                   0100  -  13                         Revision 0
                                                                  January 1995

-------
                                   TABLE  1

       SENSITIVITY  (ppb,  v/v)  OF  SAMPLING  AND ANALYSIS  FOR ALDEHYDES AND
              KETONES  IN  AMBIENT  AIR USING AN ADSORBENT CARTRIDGE
                          FOLLOWED BY GRADIENT HPLC"
Compound
Acetaldehyde
Acetone
Acrolein
Benzaldehyde
Butyraldehyde
Crotonaldehyde
2,5-Dimethyl-
    benzaldehyde
Formaldehyde
Hexanal
Isovaleraldehyde
Propionaldehyde
m-Tolualdehyde
o-Tolualdehyde
p-Tolualdehyde
Valeraldehyde
10
                              Sample Volume (L)k
20    30
40
50   100   200   300   400   500
1.36
1.28
1.29
1.07
1.21
1.22
0.68
0.64
0.65
0.53
0.61
0.61
0.45
0.43
0.43
0.36
0.40
0.41
0.34
0.32
0.32
0.27
0.30
0.31
0.27
0.26
0.26
0.21
0.24
0.24
0.
0.
0.
0.
0.
0.
14
13
13
11
12
12
0.07
0.06
0.06
0.05
0.06
0.06
0.05
0.04
0.04
0.04
0.04
0.04
0
0
0
0
0
0
.03
.03
.03
.03
.03
.03
0.03
0.03
0.03
0.02
0.02
0.02
0.97
1.45
1.09
1.15
1.28
1.02
1.02
1.02
1.15
0.49
0.73
0.55
0.57
0.64
0.51
0.51
0.51
0.57
0.32
0.48
0.36
0.38
0.43
0.34
0.34
0.34
0.38
0.24
0.36
0.27
0.29
0.32
0.25
0.25
0.25
0.29
0
0
0
0
0
0
0
0
0
.19
.29
.22
.23
.26
.20
.20
.20
.23
0.10
0.15
0.11
0.11
0.13
0.10
0.10
0.10
0.11
0.05
0.07
0.05
0.06
0.06
0.05
0.05
0.05
0.06
0.
0.
0.
0.
0.
0.
0.
0.
0.
03
05
04
04
04
03
03
03
04
0.02
0.04
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.02
0.03
0.02
0.02
0.03
0.02
0.02
0.02
0.02
      The ppb values  are  measured at 1 atm and 25°C.  The  sample  cartridge is
      eluted with 5 mL acetonitrile  and 25 p.1  is  injected  into the HPLC.   The
      maximum sampling flow through a DNPH-coated Sep-PAK is about 1.5 L/minute.
   b  A  sample  volume  of 1000  L was  also performed.   The
      sensitivity of 0.01 ppb for all the target analytes.
                                            results show  a
                                   0100  -  14
                                                 Revision 0
                                               January 1995

-------
                              FIGURE 1

           CARTRIDGE SAMPLES OF A FORMALDEHYDE  AIR STREAM
                   WITH (A) AND WITHOUT (B)  OZONE
8
            x 0
                          I
                         4           6

                                Tlmt, mln
                                                                B
10
                        x * unknown
                        0-DNPH
                        1 » formaldehyde
                        2 * acetaldehyde
                              0100 - 15
   Revision 0
 January 1995

-------
                              FIGURE 2

         SYRINGE RACKS FOR COATING AND DRYING SAMPLE CARTRIDGES
10 ml Glass
 Syringe
Adsorbent
  Tube
              (a) RACK FOR COATIN"  CARTRIDGES
Test Tube
  Rack
                                                          Waste
                                                          Beaker
     Luer-Lok
      Fitting "~
                                           «_ «_N Gas Stream
                                              Adsorbent
                                                Tubes
Waste
ViaJ




^^B

^1^

u


         (b) RACK FOR  DRYING DNPH-COATED CARTRIDGES
                              0100  - 16
   Revision 0
 January  1995

-------
                                 ATTACHMENT A
      This method does not contain a procedure for  spiking  cartridges  for  blank
spikes and matrix spikes to determine percent recovery.  Two suggested techniques
for spiking cartridges are as follows:


1)    A  spike may  be  performed  by  introducing  an  aliquot  of  a   solution
      containing  the  target  analytes  by  pipet  or  syringe  directly  onto  a
      cartridge in the field or in the laboratory.  Standard spike  and recovery
      procedures are  followed and  the  field  spike sample  is  returned to  the
      laboratory  for  analysis.   An  aliquot  of the  field spike  standard  is
      retained in the laboratory  for derivatization and comparative analysis.

2)    Another technique would include spiking  the sampling cartridge using a TGM
      555 analyzer which produces gaseous formaldehyde standards.   However,  it
      should be noted  that  the procedures required to produce accurate, dynamic,
      low-level  standard  mixtures  of  organics in  air are non-routine.    The
      techniques developed for use  in evaluating other air sampling procedures
      employ a 3-stage dynamic gas dilution  system  coupled  with  a constant-rate
      vapor generation  assembly  containing  a trioxane permeation tube  (VICI
      Medtronics Dynacal permeation device  or equivalent)  that  is maintained at
      55°C.   Trioxane vapor  is  converted stoichiometrically  to  formaldehyde
      vapor  using  a   special  high-temperature   (160"C)   catalytic   converter
      assembly.  This method of sample  introduction has been used  when  testing
      continuous sampling  apparatus.
                                   0100  -  17                         Revision 0
* U.S. GOVERNMENT PRINTING OFFICE:! 995-386-824/33253                                     JanU3ry  1995

-------