SW8461B
             METHOD STATUS TABLE
SW-846, THIRD EDITION, UPDATES I, II, AND HA
                 September 1994
   •  Use this table as a reference  guide to  identify the
      promulgation status of SW-846 methods.

   •  The methods in this table are  listed sequentially by
      number.

   •  This table should not be used as a Table of Contents for
      SW-846. Refer to the Table of Contents found in Final
      Update II (dated September 1994) for the order in which
      the methods appear in SW-846.

-------
SW-846 METHOD STATUS TABLE
      September 1994
METH NO.
THIRD EO
DATED
9/86
0010
0020
0030
1010
1020
1110
1310
"

METH NO.
FINAL
UPDATE I
DATED
7/92
~ ™*
~ ~
— •*
— •"
1020A
— w
1310A
1311
~ "*
METH NO.
FINAL
UPDT. II
DATED
9/94
" ""
•w _
"• ""
~ ""
— ~
~ ~

"
1312
METHOD TITLE
Modified Method 5
Sampling Train
Source Assessment
Sampling System
(SASS)
Volatile Organic
Sampling Train
Pensky-Martens
Closed-Cup Method
for Determining
Igm'tability
Setaflash Closed-Cup
Method for
Determining
Ignitability
Corrosivity Toward
Steel
Extraction Procedure
(EP) Toxicity Test
Method and
Structural Integrity
Test
Toxicity
Characteristic
Leaching Procedure
Synthetic
Precipitation
Leaching Procedure
SH-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol II
Chap 10
Vol II
Chap 10
Vol II
Chap 10
Vol 1C
Chap 8
Sec 8.1
Vol 1C
Chap 8
Sec 8.1
Vol 1C
Chap 8
Sec 8.2
Vol 1C
Chap 8
Sec 8.4
Vol 1C
Chap 8
Sec 8.4
Vol 1C
Chap 6
CURRENT
PROMUL-
GATED
METHOD
0010
Rev 0
9/86
0020
Rev 0
9/86
0030
Rev 0
9/86
1010
Rev 0
9/86
1020A
Rev 1
7/92
1110
Rev 0
9/86
1310A
Rev 1
7/92
1311
Rev 0
7/92
1312
Rev 0
9/94

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
KETH NO.
THIRD ED
DATED
9/86
1320
1330
3005
3010
~ ~
3020
3040
3050
METH NO.
FINAL
UPDATE I
DATED
7/92
"
1330A
3005A
3010A
— —
3020A
"
3050A
METH NO.
FINAL
UPDT. II
DATED
9/94
~ —
~ —


3015

"* ~
"
METHOD TITLE
Multiple Extraction
Procedure
Extraction Procedure
for Oily Wastes
Acid Digestion of
Waters for Total
Recoverable or
Dissolved Metals for
Analysis by FLAA or
ICP Spectroscopy
Acid Digestion of
Aqueous Samples and
Extracts for Total
Metals for Analysis
by FLAA or ICP
Spectroscopy
Microwave Assisted
Acid Digestion of
Aqueous Samples and
Extracts
Acid Digestion of
Aqueous Samples and
Extracts for Total
Metals for Analysis
by GFAA Spectroscopy
Dissolution
Procedure for Oils,
Greases, or Waxes
Acid Digestion of
Sediments, Sludges,
and Soils
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
Vol IA
Chap 3
Sec 3.2
CURRENT
PROMUL-
GATED
METHOD
1320
Rev 0
9/86
1330A
Rev 1
7/92
3005A
Rev 1
7/92
3010A
Rev 1
7/92
3015
Rev 0
9/94
3020A
Rev 1
7/92
3040
Rev 0
9/86
3050A
Rev 1
7/92

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
~ ~
3500
3510
3520
3540
~ —
3550
3580
3600
HETH NO.
FINAL
UPDATE I
DATED
7/92
*" "~
3500A
3510A
3520A
3540A
_ -
_ _
3580A
3600A
METH NO.
FINAL
UPDT. II
DATED
9/94
3051
— ~
3510B
3520B
3540B
3541
3550A
~ ~
3600B
METHOD TITLE
Microwave Assisted
Acid Digestion of
Sediments, Sludges,
Soils, and Oils
Organic Extraction
and Sample
Preparation
Separatory Funnel
Liquid-Liquid
Extraction
Continuous Liquid-
Liquid Extraction
Soxhlet Extraction
Automated Soxhlet
Extraction
Ultrasonic Extrac-
tion
Waste Dilution
Cleanup
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.2
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.2
CURRENT
PROMUL-
GATED
METHOD
3051
Rev 0
9/94
3500A
Rev 1
7/92
3510B
Rev 2
9/94
3520B
Rev 2
9/94
3540B
Rev 2
9/94
3541
Rev 0
9/94
3550A
Rev 1
9/94
3580A
Rev 1
7/92
3600B
Rev 2
9/94

-------
SW-846 METHOD STATUS TABLE (9/94),  CONTINUED
NETH NO.
THIRD ED
DATED
9/86
3610
3611
3620
3630
3640
3650
3660
"
3810
NETH NO.
FINAL
UPDATE I
DATED
7/92
3610A
3611A
3620A
3630A
~ *
3650A
3660A
• —
"
NETH NO.
FINAL
UPDT. II
DATED
9/94
"
~ ™
~ ~
3630B
3640A
"
~ ~
3665
"
NETHOD TITLE
Alumina Column
Cleanup
Alumina Column
Cleanup and
Separation of
Petroleum Hastes
Florisil Column
Cleanup
Silica Gel Cleanup
Gel -Permeation
Cleanup
Acid-Base Partition
Cleanup
Sulfur Cleanup
Sulfuric
Acid/Permanganate
Cleanup
Headspace
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec
4.2.2
Vol IB
Chap 4
Sec 4.4
CURRENT
PROMUL-
GATED
NETHOD
3610A
Rev 1
7/92
3611A
Rev 1
7/92
3620A
Rev 1
7/92
3630B
Rev 2
9/94
3640A
Rev 1
9/94
3650A
Rev 1
7/92
3660A
Rev 1
7/92
3665
Rev 0
9/94
3810
Rev 0
9/86

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
3820

5030
5040

"
6010
METH NO.
FINAL
UPDATE I
DATED
7/92
— -

5030A


— V
6010A
METH NO.
FINAL
UPDT. II
DATED
9/94
~ ~
4010
(Update
IIA,
dated
8/93)
" ~
5040A
5041
5050
"" ~
METHOD TITLE
Hexadecane
Extraction and
Screening of
Purgeable Organics
Screening for
Pentachlorophenol
by Immunoassay
Purge-and-Trap
Analysis of Sorbent
Cartridges from
Volatile Organic
Sampling Train
(VOST): Gas
Chromatography/Mass
Spectrometry
Technique
Protocol for
Analysis of Sorbent
Cartridges from
Volatile Organic
Sampling Train
(VOST): Wide -bore
Capillary Column
Technique
Bomb Preparation
Method for Solid
Waste
Inductively Coupled
Plasma- Atomic
Emission
Spectroscopy
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec 4.4
Vol IB
Chap 4
Sec 4.4
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol IB
Chap 4
Sec
4.2.1
Vol 1C
Chap 5
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
3820
Rev 0
9/86
4010
Rev 0
8/93
5030A
Rev 1
7/92
5040A
Rev 1
9/94
5041
Rev 0
9/94
5050
Rev 0
9/94
6010A
Rev 1
7/92

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
~ ™*
7000
7020
7040
7041
7060
7061
~ ~ '
7080
METH NO.
FINAL
UPDATE I
DATED
7/92
~ —
7000A
~ "
~ "*
"
"
7061A
™" ~
"
METH NO.
FINAL
UPDT. II
DATED
9/94
6020
_ •
~ *
~ —
"
7060A
™" ~
7062
7080A
METHOD TITLE
Inductively Coupled
Plasma - Mass
Spectrometry
Atomic Absorption
Methods
Aluminum (Atomic
Absorption, Direct
Aspiration)
Antimony (Atomic
Absorption, Direct
Aspiration)
Antimony (Atomic
Absorption, Furnace
Technique)
Arsenic (Atomic
Absorption, Furnace
Technique)
Arsenic (Atomic
Absorption, Gaseous
Hydride)
Antimony and Arsenic
(Atomic Absorption,
Borohydride
Reduction)
Barium (Atomic
Absorption, Direct
Aspiration)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
6020
Rev 0
9/94
7000A
Rev 1
7/92
7020
Rev 0
9/86
7040
Rev 0
9/86
7041
Rev 0
9/86
7060A
Rev 1
9/94
7061A
Rev 1
7/92
7062
Rev 0
9/94
7080A
Rev 1
9/94

-------
SW-846 METHOD STATUS TABLE (9/94),  CONTINUED
NETH NO.
THIRD ED
DATED
9/86
— -
7090
7091
7130
7131
7140
7190
7191
7195
NETH NO.
FINAL
UPDATE I
DATED
7/92
7081
— —
~ •*
~ ~
~ •"
~ *~
_ _
~ ~
"
NETH NO.
FINAL
UPDT. II
DATED
9/94
_ _
~ *
~ ~*
~ **
7131A
~ ~~
_ _
•" *
"
NETHOD TITLE
Barium (Atomic
Absorption, Furnace
Technique)
Beryllium (Atomic
Absorption, Direct
Aspiration)
Beryllium (Atomic
Absorption, Furnace
Technique)
Cadmium (Atomic
Absorption, Direct
Aspiration)
Cadmium (Atomic
Absorption, Furnace
Technique)
Calcium (Atomic
Absorption, Direct
Aspiration)
Chromium (Atomic
Absorption, Direct
Aspiration)
Chromium (Atomic
Absorption, Furnace
Technique)
Chromium, Hexavalent
(Coprecipitation)
SW-846
VOLUNE/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
NETHOD
7081
Rev 0
7/92
7090
Rev 0
9/86
7091
Rev 0
9/86
7130
Rev 0
9/86
7131A
Rev 1
9/94
7140
Rev 0
9/86
7190
Rev 0
9/86
7191
Rev 0
9/86
7195
Rev 0
9/86

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
7196
7197
7198
7200
7201
7210
•"* *~
7380
~ °~
NETH NO.
FINAL
UPDATE I
DATED
7/92
7196A
"
"• ~
"
•" *•
•. m.
7211
™ ~
7381
NETH NO.
FINAL
UPDT. II
DATED
9/94
"
"
— ••
"
•" ~"
* "
"
"

NETHOD TITLE
Chromium, Hexavalent
(Colorimetric)
Chromium, Hexavalent
(Chelation/Extrac-
tion)
Chromium, Hexavalent
(Differential Pulse
Polarography)
Cobalt (Atomic
Absorption, Direct
Aspiration)
Cobalt (Atomic
Absorption, Furnace
Technique)
Copper (Atomic
Absorption, Direct
Aspiration)
Copper (Atomic
Absorption, Furnace
Technique)
Iron (Atomic
Absorption, Direct
Aspiration)
Iron (Atomic
Absorption, Furnace
Technique)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
NETHOD
7196A
Rev 1
7/92
7197
Rev 0
9/86
7198
Rev 0
9/86
7200
Rev 0
9/86
7201
Rev 0
9/86
7210
Rev 0
9/86
7211
Rev 0
7/92
7380
Rev 0
9/86
7381
Rev 0
7/92

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
7420
7421
— —
7450
7460
"• -*
7470
7471
7480
NETH NO.
FINAL
UPDATE I
DATED
7/92
~ —
~ "•
7430
— —
— ~
7461
"
"™ *

NETH NO.
FINAL
UPDT. II
DATED
9/94
~ ~
w —
_ _
— ~
_ _
™ ~
7470A
7471A

NETHOD TITLE
Lead (Atomic
Absorption, Direct
Aspiration)
Lead (Atomic
Absorption, Furnace
Technique)
Lithium (Atomic
Absorption, Direct
Aspiration)
Magnesium (Atomic
Absorption, Direct
Aspiration)
Manganese (Atomic
Absorption, Direct
Aspiration)
Manganese (Atomic
Absorption, Furnace
Technique)
Mercury in Liquid
Waste (Manual Cold-
Vapor Technique)
Mercury in Solid or
Semi sol id Waste
(Manual Cold-Vapor
Technique)
Molybdenum (Atomic
Absorption, Direct
Aspiration)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol I A
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
NETHOD
7420
Rev 0
9/86
7421
Rev 0
9/86
7430
Rev 0
7/92
7450
Rev 0
9/86
7460
Rev 0
9/86
7461
Rev 0
7/92
7470A
Rev 1
9/94
7471A
Rev 1
9/94
7480
Rev 0
9/86

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
7481
7520
7550
7610
7740
7741
•* —
7760
"" —
METH NO.
FINAL
UPDATE I
DATED
7/92
"
"
~ ' ~
"
_ _
"
"• ~
7760A
7761
METH NO.
FINAL
UPDT. II
DATED
9/94
~ ~
_ •
*• ~
"
~ ~
7741A
7742
~ •*
— —
METHOD TITLE
Molybdenum (Atomic
Absorption, Furnace
Technique)
Nickel (Atomic
Absorption, Direct
Aspiration)
Osmium (Atomic
Absorption, Direct
Aspiration)
Potassium (Atomic
Absorption, Direct
Aspiration)
Selenium (Atomic
Absorption, Furnace
Technique)
Selenium (Atomic
Absorption, Gaseous
Hydride)
Selenium (Atomic
Absorption,
Borohydride
Reduction)
Silver (Atomic
Absorption, Direct
Aspiration)
Silver (Atomic
Absorption, Furnace
Technique)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
METHOD
7481
Rev 0
9/86
7520
Rev 0
9/86
7550
Rev 0
9/86
7610
Rev 0
9/86
7740
Rev 0
9/86
7741A
Rev 1
9/94
7742
Rev 0
9/94
7760A
Rev 1
7/92
7761
Rev 0
7/92
                                      10

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
7770
~ "•
7840
7841
7870
7910
7911
7950
"
NETH NO.
FINAL
UPDATE I
DATED
7/92
~ **
7780
~ —
*~ "*
"
— ™
~ "
"
7951
NETH NO.
FINAL
UPDT. II
DATED
9/94
*" ~
" ~
<. «
~ ~
"
~ ~
~ ~
"

NETHOD TITLE
Sodium (Atomic
Absorption, Direct
Aspiration)
Strontium (Atomic
Absorption, Direct
Aspiration)
Thallium (Atomic
Absorption, Direct
Aspiration)
Thallium (Atomic
Absorption, Furnace
Technique)
Tin (Atomic
Absorption, Direct
Aspiration)
Vanadium (Atomic
Absorption, Direct
Aspiration)
Vanadium (Atomic
Absorption, Furnace
Technique)
Zinc (Atomic
Absorption, Direct
Aspiration)
Zinc (Atomic
Absorption, Furnace
Technique)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
Vol IA
Chap 3
Sec 3.3
CURRENT
PROMUL-
GATED
NETHOD
7770
Rev 0
9/86
7780
Rev 0
7/92
7840
Rev 0
9/86
7841
Rev 0
9/86
7870
Rev 0
9/86
7910
Rev 0
9/86
7911
Rev 0
9/86
7950
Rev 0
9/86
7951
Rev 0
7/92
                                      11

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
NETH NO.
THIRD ED
DATED
9/86
8000
8010

8015
8020

8030
"
NETH NO.
FINAL
UPDATE I
DATED
7/92
8000A
8010A
8011
80 ISA
~ •
8021
8030A
— *~
METH NO.
FINAL
UPDT. II
DATED
9/94
"
8010B

~ ~
8020A
8021A
. _
8031
METHOD TITLE
Gas Chromatography
Halogenated Volatile
Organics by Gas
Chromatography
1,2-Dibromoethane
and l,2-Dibromo-3-
chloropropane by
Microextraction and
Gas Chromatography
Nonhalogenated
Volatile Organics by
Gas Chromatography
Aromatic Volatile
Organics by Gas
Chromatography
Halogenated
Volatiles by Gas
Chromatography Using
Photoionization and
Electrolytic
Conductivity
Detectors in Series:
Capillary Column
Technique
Acrolein and
Acrylonitrile by Gas
Chromatography
Acrylonitrile by Gas
Chromatography
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
CURRENT
PROMUL-
GATED
METHOD
8000A
Rev 1
7/92
8010B
Rev 2
9/94
8011
Rev 0
7/92
8015A
Rev 1
7/92
8020A
Rev 1
9/94
8021A
Rev 1
9/94
8030A
Rev 1
7/92
8031
Rev 0
9/94
                                      12

-------
SH-846 METHOD STATUS TABLE (9/94), CONTINUED
HETH NO.
THIRD ED
DATED
9/86
~ ~
8040
8060

~ *"
8080

8090
HETH NO.
FINAL
UPDATE I
DATED
7/92
~ ~
8040A
~ ~

8070


~ ™ '
NETH NO.
FINAL
UPDT. II
DATED
9/94
8032
"
"™ ~
8061
— ~
8080A
8081
"
METHOD TITLE
Acryl amide by Gas
Chromatography
Phenols by Gas
Chromatography
Phthalate Esters
Phthalate Esters by
Capillary Gas
Chromatography with
Electron Capture
Detection (GC/ECD)
Nitrosamines by Gas
Chromatography
Organochlorine Pes-
ticides and
Polychlorinated
Biphenyls by Gas
Chromatography
Organochlorine
Pesticides and PCBs
as Aroclors by Gas
Chromatography:
Capillary Column
Technique
Nitroaromatics and
Cyclic Ketones
SH-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
CURRENT
PROMUL-
GATED
METHOD
8032
Rev 0
9/94
8040A
Rev 1
7/92
8060
Rev 0
9/86
8061
Rev 0
9/94
8070
Rev 0
7/92
8080A
Rev 1
9/94
8081
Rev 0
9/94
8090
Rev 0
9/86
                                      13

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
8100
"" ~
8120

8140

8150

METH NO.
FINAL
UPDATE I
DATED
7/92
"
8110
"* ~

— —
8141
8150A

METH NO.
FINAL
UPDT. II
DATED
9/94
_ *
"
8120A
8121
— ~
8141A
8150B
8151
METHOD TITLE
Polynuclear Aromatic
Hydrocarbons
Haloethers by Gas
Chromatography
Chlorinated
Hydrocarbons by Gas
Chromatography
Chlorinated
Hydrocarbons by Gas
Chromatography:
Capillary Column
Technique
Organophosphorus
Pesticides
Organophosphorus
Compounds by Gas
Chromatography:
Capillary Column
Technique
Chlorinated
Herbicides by Gas
Chromatography
Chlorinated
Herbicides by GC
Using Methylation or
Pentaf 1 uorobenzyl -
ation Derivati-
zation: Capillary
Column Technique
SM-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
Vol IB
Chap 4
Sec
4.3.1
CURRENT
PROMUL-
GATED
METHOD
8100
Rev 0
9/86
8110
Rev 0
7/92
8120A
Rev 1
9/94
8121
Rev 0
9/94
8140
Rev 0
9/86
8141A
Rev 1
9/94
8150B
Rev 2
9/94
8151
Rev 0
9/94
                                      14

-------
SW-846 METHOD STATUS TABLE (9/94),  CONTINUED
HETH NO.
THIRD ED
DATED
9/86
8240
8250

8270

8280
METH NO.
FINAL
UPDATE I
DATED
7/92
8240A

8260
8270A


METH NO.
FINAL
UPDT. II
DATED
9/94
8240B
8250A
8260A
8270B
8275

METHOD TITLE
Volatile Organic
Compounds by Gas
Chromatography/Mass
Spectrometry (GC/MS)
Semi volatile Organic
Compounds
by Gas
Chromatography/Mass
Spectrometry (GC/MS)
Volatile Organic
Compounds by Gas
Chromatography/Mass
Spectrometry
(GC/MS): Capillary
Column Technique
Semivolatile Organic
Compounds by Gas
Chromatography/Mass
Spectrometry
(GC/MS): Capillary
Column Technique
Thermal
Chromatography/Mass
Spectrometry (TC/MS)
for Screening
Semivolatile Organic
Compounds
The Analysis of
Polychlorinated
Dibenzo-p-Dioxins
and Polychlorinated
Dibenzofurans
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec 4.4
Vol IB
Chap 4
Sec
4.3.2
CURRENT
PROMUL-
GATED
METHOD
8240B
Rev 2
9/94
8250A
Rev 1
9/94
8260A
Rev 1
9/94
8270B
Rev 2
9/94
8275
Rev 0
9/94
8280
Rev 0
9/86
                                      15

-------
SW-846 METHOD STATUS TABLE (9/94),  CONTINUED
METH NO.
THIRD ED
DATED
9/86

8310


"
METH NO.
FINAL
UPDATE I
DATED
7/92

" ~


"
METH NO.
FINAL
UPDT. II
DATED
9/94
8290
— ~
8315
8316
8318
METHOD TITLE
Polychlorinated
Dibenzodioxins
(PCDDs) and
Polychlorinated
Dibenzofurans
(PCDFs) by High-
Resolution Gas
Chromatography/High-
Resolution Mass
Spectrometry
(HRGC/HRMS)
Polynuclear Aromatic
Hydrocarbons
Determination of
Carbonyl Compounds
by High Performance
Liquid
Chromatography
(HPLC)
Aery 1 amide,
Acrylonitrile and
Acrolein by High
Performance Liquid
Chromatography
(HPLC)
N-Methyl carbamates
by High Performance
Liquid Chroma-
tography (HPLC)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.2
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
CURRENT
PROMUL-
GATED
METHOD
8290
Rev 0
9/94
8310
Rev 0
9/86
8315
Rev 0
9/94
8316
Rev 0
9/94
8318
Rev 0
9/94
                                      16

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86




9010
9012
METH NO.
FINAL
UPDATE I
DATED
7/92




9010A
"
METH NO.
FINAL
UPDT. II
DATED
9/94
8321
8330
8331
8410
** ~
"
METHOD TITLE
Solvent Extractable
Non-Volatile
Compounds by High
Performance Liquid
Chromatography/Ther-
mospray/Mass
Spectrometry
(HPLC/TSP/MS) or
Ultraviolet (UV)
Detection
Nitroaromatics and
Nitramines by High
Performance Liquid
Chromatography
(HPLC)
Tetrazene by Reverse
Phase High
Performance Liquid
Chromatography
(HPLC)
Gas Chroma-
tography/Fourier
Transform Infrared
(GC/FT-IR) Spec-
trometry for
Semi volatile
Organics: Capillary
Column
Total and Amenable
Cyanide
(Colorimetric,
Manual)
Total and Amenable
Cyanide
(Colorimetric,
Automated UV)
SM-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.3
Vol IB
Chap 4
Sec
4.3.4
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
8321
Rev 0
9/94
8330
Rev 0
9/94
8331
Rev 0
9/94
8410
Rev 0
9/94
9010A
Rev 1
7/92
9012
Rev 0
9/86
                                      17

-------
SW-846 METHOD STATUS TABLE (9/94),  CONTINUED
METH NO.
THIRD ED
DATED
9/86
"~ ~
9020
"
9022
9030
"
9035
9036
9038
METH NO.
FINAL
UPDATE I
DATED
7/92
9013
9020A
9021
^ ~
9030A
9031
"

"
METH NO.
FINAL
UPDT. II
DATED
9/94
— ~
90206
~ ~
"
~ ~
*" ~
"


METHOD TITLE
Cyanide Extraction
Procedure for Sol ids
and Oils
Total Organic
Hal ides (TOX)
Purgeable Organic
Hal ides (POX)
Total Organic
Hal ides (TOX) by
Neutron Activation
Analysis
Acid-Soluble and
Acid-Insoluble
Sulfides
Extractable Sulfides
Sulfate
(Colorimetric,
Automated,
Chloranilate)
Sulfate
(Colorimetric,
Automated,
Methyl thymol Blue,
AA II)
Sulfate
(Turbidimetric)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
9013
Rev 0
7/92
9020B
Rev 2
9/94
9021
Rev 0
7/92
9022
Rev 0
9/86
9030A
Rev 1
7/92
9031
Rev 0
7/92
9035
Rev 0
9/86
9036
Rev 0
9/86
9038
Rev 0
9/86
                                      18

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9040
9041
9045
9050
>. ™
9060
9065
9066
9067
METH NO.
FINAL
UPDATE I
DATED
7/92
~ •*
9041A
9045A
*• *•
— —
*• "•

"

NETH NO.
FINAL
UPDT. II
DATED
9/94
9040A
"" ™
9045B
"* *
9056
~" "
— —
"
"
METHOD TITLE
pH Electrometric
Measurement
pH Paper Method
Soil and Waste pH
Specific Conductance
Determination of
Inorganic Anions by
Ion Chromatography
Total Organic Carbon
Phenol ics
(Spectrophotometri c ,
Manual 4-AAP with
Distillation)
Phenol ics
(Colorimetric,
Automated 4-AAP with
Distillation)
Phenol ics
(Spectrophotometri c,
MBTH with
Distillation)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
9040A
Rev 1
9/94
9041A
Rev 1
7/92
9045B
Rev 2
9/94
9050
Rev 0
9/86
9056
Rev 0
9/94
9060
Rev 0
9/86
9065
Rev 0
9/86
9066
Rev 0
9/86
9067
Rev 0
9/86
                                      19

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
HETH NO.
THIRD ED
DATED
9/86
9070
9071



9080
9081
HETH NO.
FINAL
UPDATE I
DATED
7/92





"
"
METH NO.
FINAL
UPDT. II
DATED
9/94

9071A
9075
9076
9077
"
"
METHOD TITLE
Total Recoverable
Oil & Grease
(Gravimetric,
Separatory Funnel
Extraction)
Oil and Grease
Extraction Method
for Sludge and
Sediment
Samples
Test Method for
Total Chlorine in
New and Used
Petroleum Products
by X-Ray
Fluorescence
Spectrometry (XRF)
Test Method for
Total Chlorine in
New and Used
Petroleum Products
by Oxidative
Combustion and
Microcoulometry
Test Methods for
Total Chlorine in
New and Used
Petroleum Products
(Field Test Kit
Methods)
Cation-Exchange
Capacity of Soils
(Ammonium Acetate)
Cation-Exchange
Capacity of Soils
(Sodium Acetate)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 6
Vol 1C
Chap 6
CURRENT
PROMUL-
GATED
METHOD
9070
Rev 0
9/86
9071A
Rev 1
9/94
9075
Rev 0
9/94
9076
Rev 0
9/94
9077
Rev 0
9/94
9080
Rev 0
9/86
9081
Rev 0
9/86
                                      20

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9090
9095
*~ ~
9100
9131
9132
9200
9250
9251
METH NO.
FINAL
UPDATE I
DATED
7/92
9090A
_ -
~ ~"

"
™* *"

*" ~

METH NO.
FINAL
UPDT. II
DATED
9/94
~ "
~ "
9096

~ ""
— ••
~ "
~ *
"* *"
METHOD TITLE
Compatibility Test
for Wastes and
Membrane Liners
Paint Filter Liquids
Test
Liquid Release Test
(LRT) Procedure
Saturated Hydraulic
Conductivity,
Saturated Leachate
Conductivity, and
Intrinsic
Permeability
Total Col i form:
Multiple Tube
Fermentation
Technique
Total Col i form:
Membrane Filter
Technique
Nitrate
Chloride
(Colorimetric,
Automated
Ferricyanide AAI)
Chloride
(Colorimetric,
Automated
Ferricyanide AAI I)
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 5
CURRENT
PROMUL-
GATED
METHOD
9090A
Rev 1
7/92
9095
Rev 0
9/86
9096
Rev 0
9/94
9100
Rev 0
9/86
9131
Rev 0
9/86
9132
Rev 0
9/86
9200
Rev 0
9/86
9250
Rev 0
9/86
9251
Rev 0
9'S6
                                      21

-------
SW-846 METHOD STATUS TABLE (9/94), CONTINUED
METH NO.
THIRD ED
DATED
9/86
9252
~ *
9310
9315
9320
HCN Test
Method
H2S Test
Method
METH NO.
FINAL
UPDATE I
DATED
7/92
•" ~
*™ *"
— —
— ••
~ —
HCN Test
Method
H2S Test
Method
METH NO.
FINAL
UPDT. II
DATED
9/94
9252A
9253
™* ~
— ~
"~ ~
HCN Test
Method
H2S Test
Method
METHOD TITLE
Chloride
(Titrimetric,
Mercuric Nitrate)
Chloride
(Titrimetric, Silver
Nitrate)
Gross Alpha and
Gross Beta
Alpha- Emitting
Radium Isotopes
Radium-228
Test Method to
Determine Hydrogen
Cyanide Released
from Wastes
Test Method to
Determine Hydrogen
Sulfide Released
from Wastes
SW-846
VOLUME/
CHAPTER/
SECTION
LOCATION
Vol 1C
Chap 5
Vol 1C
Chap 5
Vol 1C
Chap 6
Vol 1C
Chap 6
Vol 1C
Chap 5
Vol 1C
Chap 7
Sec 7.3
Vol 1C
Chap 7
Sec 7.3
CURRENT
PROMUL-
GATED
METHOD
9252A
Rev 1
9/94
9253
Rev 0
9/94
9310
Rev 0
9/86
9315
Rev 0
9/86
9320
Rev 0
9/86
Guidance
Method
Only
Guidance
Method
Only
                                      22

-------
VOLUME  ONE,

     SECTION B
                        Revision      0
                        Date  September 1986

-------
For Rale by the Superintendent of Documents, U.S. Government Printing Office Washington, D.C. 20402

-------
                                  DISCLAIMER


      Mention  of  trade  names  or  commercial  products  does  not  constitute
endorsement  or recommendation  for  use  by the U.S.  Environmental  Protection
Agency.

      SW-846 methods are designed to be used with equipment from any manufacturer
that results In suitable method performance (as assessed by accuracy, precision,
detection limits and matrix compatibility).  In several SW-846 methods, equipment
specifications  and  settings  are  given for the specific instrument used during
method development, or  subsequently  approved for use  in the  method.   These
references are made to provide the best possible guidance  to  laboratories using
this manual.   Equipment  not  specified  in  the  method may be used as long as the
laboratory  achieves equivalent or superior method performance.   If alternate
equipment is used, the laboratory must follow the manufacturer's instructions for
their particular instrument.

      Since  many  types  and  sizes  of glassware and  supplies are commercially
available, and  since  it is  possible  to prepare reagents and standards in many
different ways, those specified in these methods may  be  replaced by any similar
types as long  as this  substitution  does not affect the overall quality of the
analyses.
                                 DISCLAIMER  -  1                      Revision 0
                                                                     July 1992

-------
                                   ABSTRACT

Test Methods  for Evaluating Solid Waste, Physical/Chemical  Methods  (SW-846)
provides test procedures and guidance which are recommended for use in conducting
the evaluations and measurements needed to comply with the Resource Conservation
and  Recovery  Act  (RCRA), Public  Law 94-580, as  amended.    These  methods are
approved  by the  U.S.  Environmental  Protection  Agency  for  obtaining  data to
satisfy the requirements  of  40 CFR Parts 122  through  270 promulgated under RCRA,
as amended.   This manual presents the  state-of-the-art  in routine analytical
tested  adapted for the  RCRA  program.    It  contains procedures  for  field and
laboratory  quality control, sampling,  determining   hazardous  constituents in
wastes,  determining   the   hazardous   characteristics  of  wastes  (toxicity,
ignitability,  reactivity,   and   corrosivity),   and   for  determining  physical
properties of  wastes.  It also contains  guidance on how to select appropriate
methods.

      Several of the hazardous waste regulations  under Subtitle C of RCRA require
that  specific testing methods  described in SW-846  be  employed  for  certain
applications.  Refer to 40 Code of Federal Regulations (CFR), Parts 260 through
270, for those specific requirements. Any reliable analytical method may be used
to meet other  requirements under Subtitle C of RCRA.
                                 ABSTRACT - :                       Revision 2
                                                                September 1994

-------
                           TABLE  OF  CONTENTS
                                     VOLUME ONE

                                      SECTION A
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE
ACKNOWLEDGEMENTS
                    PART I    METHODS FOR ANALYTES AND PROPERTIES

CHAPTER ONE -- QUALITY CONTROL
1.0
2.0
3.0
4.0
5.0
6.0
Introduction
QA Project Plan
Field Operations
Laboratory Operations
Definitions
References
CHAPTER TWO -- CHOOSING THE CORRECT PROCEDURE

      2.1   Purpose
      2.2   Required  Information
      2.3   Implementing the Guidance
      2.4   Characteristics
      2.5   Ground Water
      2.6   References

CHAPTER THREE -- METALLIC ANALYTES

      3.1   Sampling  Considerations
      3.2   Sample Preparation Methods

            Method 3005A:     Acid  Digestion  of Waters  for  Total  Recoverable  or
                              Dissolved Metals for Analysis by Flame Atomic Absorption
                              (FLAA) or Inductively Coupled  Plasma  (ICP) Spectroscopy
            Method 3010A:     Acid Digestion of Aqueous Samples and Extracts for Total
                              Metals for Analysis by Flame Atomic Absorption (FLAA) or
                              Inductively Coupled Plasma (ICP) Spectroscopy
            Method 3015:      Microwave Assisted Acid Digestion of Aqueous Samples and
                              Extracts
                                    CONTENTS - 1                          Revision 2
                                                                      September 1994

-------
      Method 3020A:
      Method 3040:
      Method 3050A:
      Method 3051:
           Acid Digestion of Aqueous Samples and Extracts  for Total
           Metals  for   Analysis   by  Graphite   Furnace  Atomic
           Absorption (6FAA)  Spectroscopy
           Dissolution Procedure for Oils, Greases, or Waxes
           Acid Digestion of Sediments, Sludges, and Soils
           Microwave Assisted Acid Digestion of Sediments, Sludges,
           Soils, and Oils
3.3   Methods for Determination of Metals
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
      Method
6010A:
6020:
7000A:
7020:
7040:
7041:
7060A:
7061A:
7062:
7080A:
7081:
7090:
7091:
7130:
7131 A:
7140:
7190:
7191:
7195:
7196A:
7197:
7198:
7200:
7201:
7210:
7211:
7380:
7381:
7420:
7421:
7430:
7450:
7460:
7461:
7470A:
7471A:
      Method 7480:
      Method 7481:
      Method 7520:
      Method 7550:
      Method 7610:
      Method 7740:
Inductively Coupled Plasma-Atomic Emission Spectroscopy
Inductively Coupled Plasma - Mass Spectrometry
Atomic Absorption Methods
Aluminum (AA, Direct Aspiration)
Antimony (AA, Direct Aspiration)
Antimony (AA, Furnace Technique)
Arsenic (AA, Furnace Technique)
Arsenic (AA, Gaseous Hydride)
Antimony and Arsenic (AA, Borohydride Reduction)
Barium (AA, Direct Aspiration)
Barium (AA, Furnace Technique)
Beryllium  (AA, Direct Aspiration)
Beryllium  (AA, Furnace Technique)
Cadmium (AA, Direct Aspiration)
Cadmium (AA, Furnace Technique)
Calcium (AA, Direct Aspiration)
Chromium (AA, Direct Aspiration)
Chromium (AA, Furnace Technique)
Chromium,  Hexavalent (Coprecipitation)
Chromium,  Hexavalent (Colorimetric)
Chromium,  Hexavalent (Chelation/Extraction)
Chromium,  Hexavalent (Differential Pulse Polarography)
Cobalt (AA, Direct Aspiration)
Cobalt (AA, Furnace Technique)
Copper (AA, Direct Aspiration)
Copper (AA, Furnace Technique)
Iron  (AA,  Direct Aspiration)
Iron  (AA,  Furnace Technique)
Lead  (AA,  Direct Aspiration)
Lead  (AA,  Furnace Technique)
Lithium (AA, Direct Aspiration)
Magnesium  (AA, Direct Aspiration)
Manganese  (AA, Direct Aspiration)
Manganese  (AA, Furnace Technique)
Mercury in Liquid Waste  (Manual Cold-Vapor Technique)
Mercury in Solid or Semi sol id Waste  (Manual Cold-Vapor
Technique)
Molybdenum (AA, Direct Aspiration)
Molybdenum (AA, Furnace  Technique)
Nickel (AA, Direct Aspiration)
Osmium (AA, Direct Aspiration)
Potassium  (AA, Direct Aspiration)
Selenium (AA, Furnace Technique)
                              CONTENTS - 2
                                                        Revision 2
                                                   September  1994

-------
            Method 7741A:     Selenium (AA; Gaseous Hydride)
            Method 7742:      Selenium (AA, Borohydride Reduction)
            Method 7760A:     Silver (AA, Direct Aspiration)
            Method 7761:      Silver (AA, Furnace Technique)
            Method 7770:      Sodium (AA, Direct Aspiration)
            Method 7780:      Strontium  (AA, Direct Aspiration)
            Method 7840:      Thallium (AA, Direct Aspiration)
            Method 7841:      Thallium (AA, Furnace Technique)
            Method 7870:      Tin  (AA, Direct Aspiration)
            Method 7910:      Vanadium (AA, Direct Aspiration)
            Method 7911:      Vanadium (AA, Furnace Technique)
            Method 7950:      Zinc (AA,  Direct Aspiration)
            Method 7951:      Zinc (AA,  Furnace Technique)
APPENDIX -- COMPANY REFERENCES
        NOTE;   A suffix of "A" in the method  number  indicates revision one
        (the  method  has been revised once).  A  suffix  of  "B"  in the method
        number  indicates revision two (the method has  been revised twice). In
        order to properly  document the method  used for analysis, the entire
        method  number Including the suffix  letter designation  (e.g., A or B)
        must  be Identified by the analyst.   A  method  reference found within
        the  RCRA regulations  and the text  of SW-846 methods  and  chapters
        refers  to the latest  promulgated revision of the method, even though
        the method number  does not include  the appropriate letter suffix.
                                    CONTENTS - 3
    Revision
September

-------
                                     VOLUME ONE

                                      SECTION B
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE
ACKNOWLEDGEMENTS

CHAPTER ONE. REPRINTED -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References

CHAPTER FOUR -- ORGANIC ANALYTES

      4.1   Sampling Considerations
      4.2   Sample Preparation Methods

            4.2.1       Extractions and Preparations

            Method 3500A:     Organic Extraction and Sample Preparation
            Method 3510B:     Separatory Funnel Liquid-Liquid Extraction
            Method 3520B:     Continuous Liquid-Liquid Extraction
            Method 3540B:     Soxhlet Extraction
            Method 3541:      Automated Soxhlet Extraction
            Method 3550A:     Ultrasonic Extraction
            Method 3580A:     Waste Dilution
            Method 5030A:     Purge-and-Trap
            Method 5040A:     Analysis  of  Sorbent Cartridges from  Volatile  Organic
                              Sampling   Train   (VOST):      Gas   Chromatography/Mass
                              Spectrometry Technique
            Method 5041:      Protocol  for  Analysis  of  Sorbent  Cartridges  from
                              Volatile  Organic Sampling  Train  (VOST):    Wide-bore
                              Capillary Column Technique
            Method 5100:      Determination of the Volatile Organic Concentration of
                              Waste Samples
            Method 5110:      Determination of Organic Phase Vapor Pressure in Waste
                              Samples

            4.2.2       Cleanup

            Method 3600B:     Cleanup
            Method 3610A:     Alumina Column Cleanup
                                    CONTENTS - 4                          Revision 2
                                                                      September 1994

-------
      Method 3611A:
      Method
      Method
      Method
      Method
      Method
      Method
3620A:
3630B:
3640A:
3650A:
3660A:
3665:
Alumina    Column
Petroleum Wastes
Florisil Column Cleanup
Silica Gel Cleanup
Gel-Permeation Cleanup
Acid-Base Partition Cleanup
Sulfur Cleanup
Sulfuric Acid/Permanganate Cleanup
                                Cleanup     and    Separation    of
4.3   Determination of Organic Analytes
      4.3.1
     Gas Chromatographic Methods
      Method 8000A:
      Method 8010B:
      Method 8011:

      Method 8015A:
      Method 8020A:
      Method 8021A:
      Method
      Method
      Method
      Method
      Method
      Method
8030A:
8031:
8032:
8040A:
8060:
8061:
      Method 8070:
      Method 8080A:

      Method 8081:

      Method 8090:
      Method 8100:
      Method 8110:
      Method 8120A:
      Method 8121:

      Method 8140:
      Method 8141A:

      Method 8150B:
      Method 8151:
                                                               by
Gas Chromatography
Halogenated Volatile Organics by Gas Chromatography
1,2-Dibromoethane  and  l,2-Dibromo-3-chloropropane
Microextraction and Gas Chromatography
Nonhalogenated Volatile Organics by Gas Chromatography
Aromatic Volatile Organics by Gas Chromatography
Halogenated  Volatiles   by  Gas  Chromatography  Using
Photoionization and Electrolytic Conductivity Detectors
in Series: Capillary Column Technique
Acrolein and Acrylonitrile by Gas Chromatography
Acrylonitrile by Gas Chromatography
Acrylamide by Gas Chromatography
Phenols by Gas Chromatography
Phthalate Esters
Phthalate Esters by Capillary  Gas Chromatography with
Electron Capture Detection (GC/ECD)
Nitrosamines by Gas Chromatography
Organochlorine Pesticides  and Polychlorinated Biphenyls
by Gas Chromatography
Organochlorine Pesticides and  PCBs  as  Aroclors by Gas
Chromatography:  Capillary Column Technique
Nitroaromatics and Cyclic Ketones
Polynuclear Aromatic Hydrocarbons
Haloethers by Gas Chromatography
Chlorinated Hydrocarbons by Gas Chromatography
Chlorinated   Hydrocarbons  by  Gas   Chromatography:
Capillary Column Technique
Organophosphorus Pesticides
Organophosphorus  Compounds  by   Gas  Chromatography:
Capillary Column Technique
Chlorinated Herbicides  by Gas Chromatography
Chlorinated  Herbicides by  GC Using  Methylation  or
Pentafluorobenzylation  Derivatization:  Capillary Column
Technique
                              CONTENTS -  5
                                                       Revision 2
                                                   September 1994

-------
4.3.2
Gas Chromatographic/Mass Spectroscopic Methods
Method 8240B:

Method 8250A:

Method 8260A:

Method 8270B:


Method 8280:

      Appendix A:
      Appendix B:

Method 8290:



      Appendix A:
      Volatile Organic Compounds by  Gas  Chromatography/Mass
      Spectrometry (GC/MS)
      Semivolatile     Organic     Compounds    by    Gas
      Chromatography/Mass Spectrometry (GC/MS)
      Volatile Organic Compounds by  Gas  Chromatography/Mass
      Spectrometry (GC/MS):  Capillary Column Technique
      Semivolatile     Organic     Compounds    by    Gas
      Chromatography/Mass  Spectrometry  (GC/MS):   Capillary
      Column Technique
      The Analysis of Polychlorinated  Dibenzo-p-Dioxins and
      Polychlorinated Dibenzofurans
            Signal-to-Noise  Determination Methods
            Recommended  Safety and  Handling Procedures for
            PCDDs/PCDFs
      Polychlorinated    Dibenzodioxins    (PCDDs)    and
      Polychlorinated Dibenzofurans (PCDFs) by High-Resolution
      Gas  Chromatography/High-Resolution  Mass  Spectrometry
      (HRGC/HRMS)
            Procedures   for   the    Collection,    Handling,
            Analysis, and Reporting  of Wipe  Tests  Performed
            within the Laboratory
4.3.3

Method 8310:
Method 8315:
      Appendix A:

Method 8316:

Method 8318:

Method 8321:


Method 8330:

Method 8331:
High Performance Liquid Chromatographic Methods
      Polynuclear Aromatic Hydrocarbons
      Determination of Carbonyl Compounds by High Performance
      Liquid Chromatography (HPLC)
            Recrystallization of  2,4-Dinitrophenylhydrazine
            (DNPH)
      Acrylamide,   Acrylonitrile   and   Acrolein   by   High
      Performance Liquid Chromatography (HPLC)
      N-Methylcarbamates   by   High    Performance    Liquid
      Chromatography (HPLC)
      Solvent  Extractable Non-Volatile  Compounds  by  High
      Performance   Liquid   Chromatography/Thermospray/Mass
      Spectrometry (HPLC/TSP/MS) or Ultraviolet (UV) Detection
      Nitroaromatics and Nitramines by High Performance Liquid
      Chromatography (HPLC)
      Tetrazene  by Reverse  Phase High  Performance  Liquid
      Chromatography (HPLC)
4.3.4

Method 8410:
Fourier Transform Infrared Methods

      Gas Chromatography/Fourier Transform Infrared (GC/FT-IR)
      Spectrometry  for  Semivolatile  Organics:    Capillary
      Column
                        CONTENTS - 6
                                                  Revision  2
                                              September 1994

-------
      4.4   Miscellaneous Screening Methods

            Method 3810:      Headspace
            Method 3820:      Hexadecane  Extraction  and   Screening   of  Purgeable
                              Organics
            Method 4010:      Screening for Pentachlorophenol by Immunoassay
            Method 8275:      Thermal  Chromatography/Mass  Spectrometry  (TC/MS)  for
                              Screening Semivolatile Organic Compounds

APPENDIX -- COMPANY REFERENCES
        NOTE:   A  suffix  of "A" in the method number  indicates  revision  one
        (the method has  been  revised  once).   A suffix of  "B"  in  the  method
        number  indicates revision two  (the method has been revised twice). In
        order to  properly document the method used  for  analysis,  the  entire
        method  number Including the suffix letter designation (e.g., A or B)
        must be Identified by  the analyst.   A method  reference  found  within
        the  RCRA  regulations   and the  text  of  SW-846 methods  and chapters
        refers  to the latest promulgated revision of the method, even  though
        the method number does not include the appropriate letter suffix.
                                    CGN'ENTS -  7                          Revision 2
                                                                      September 1994

-------
                         VOLUME ONE

                         SECTION C
                                                                     UV)
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE

CHAPTER ONE, REPRINTED -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References

CHAPTER FIVE -- MISCELLANEOUS TEST METHODS

            Method 5050:      Bomb Preparation Method for Solid Waste
            Method 9010A:     Total and Amenable Cyanide (Colorimetric, Manual)
            Method 9012:      Total and Amenable Cyanide (Colorimetric,  Automated
            Method 9013:      Cyanide Extraction Procedure for Solids and Oils
            Method 9020B:     Total Organic Hal ides  (TOX)
            Method 9021:      Purgeable Organic Hal ides  (POX)
            Method 9022:      Total  Organic  Hal ides  (TOX)   by   Neutron  Activation
                              Analysis
                              Acid-Soluble and Acid-Insoluble Sulfides
                              Extractable Sulfides
                              Sulfate (Colorimetric, Automated,
                              Sulfate (Colorimetric,  Automated,
                              II)
            Method 9038:      Sulfate (Turbidimetric)
            Method 9056:      Determination of Inorganic Anions by Ion Chromatography
            Method 9060:      Total Organic Carbon
            Method 9065:      Phenolics   (Spectrophotometric,
                              Distillation)
                              Phenolics    (Colorimetric,    Automated
                              Distillation)
            Method 9067:      Phenolics (Spectrophotometric,  MBTH with  Distillation)
            Method 9070:      Total Recoverable Oil & Grease (Gravimetric, Separatory
                              Funnel Extraction)
            Method 9071A:     Oil and Grease Extraction  Method for Sludge and Sediment
                              Samples
            Method 9075:      Test Method for Total Chlorine  in New and Used  Petroleum
                              Products by X-Ray Fluorescence Spectrometry (XRF)
            Method 9076:      Test Method for Total Chlorine  in New and Used  Petroleum
                              Products by Oxidative  Combustion and Microcoulometry
Method 9030A:
Method 9031:
Method 9035:
Method 9036:
 Chloranilate)
Methyl thymol Blue,
AA
Method 9066:
                                                    Manual   4-AAP  with
         4-AAP    with
                        CONTENTS  - 8
                                                                           Revision  2
                                                                       September  1994

-------
            Method 9077:

                  Method A:
                  Method B:

                  Method C:
            Method 9131:
            Method 9132:
            Method 9200:
            Method 9250:
            Method 9251:
            Method 9252A:
            Method 9253:
            Method 9320:

CHAPTER SIX -- PROPERTIES

            Method 1312:
            Method 1320:
            Method 1330A:
            Method 9040A:
            Method 9041A:
            Method 9045B:
            Method 9050:
            Method 9080:
            Method 9081:
            Method 9090A:
            Method 9095:
            Method 9096:
                  Appendix A:
            Method 9100:

            Method 9310:
            Method 9315:
                              Test  Methods  for  Total   Chlorine  in   New   and  Used
                              Petroleum Products  (Field Test Kit Methods)
                                    Fixed End Point Test Kit Method
                                    Reverse Titration Quantitative End Point Test Kit
                                    Method
                              Direct Titration Quantitative End Point  Test Kit Method
                              Total Coliform:  Multiple Tube Fermentation Technique
                              Total Coliform:  Membrane Filter Technique
                              Nitrate
                              Chloride (Colorimetric, Automated Ferricyanide AAI)
                              Chloride (Colorimetric, Automated Ferricyanide AAII)
                              Chloride (Titrimetric, Mercuric Nitrate)
                              Chloride (Titrimetric, Silver Nitrate)
                              Radium-228
                              Synthetic Precipitation Leaching Procedure
                              Multiple Extraction Procedure
                              Extraction Procedure for Oily Wastes
                              pH Electrometric Measurement
                              pH Paper Method
                              Soil and Waste pH
                              Specific Conductance
                              Cation-Exchange Capacity of Soils  (Ammonium Acetate)
                              Cation-Exchange Capacity of Soils  (Sodium Acetate)
                              Compatibility Test for Wastes and  Membrane Liners
                              Paint Filter Liquids Test
                              Liquid Release Test (LRT) Procedure
                                    LRT Pre-Test
                              Saturated  Hydraulic  Conductivity,  Saturated Leachate
                              Conductivity, and Intrinsic Permeability
                              Gross Alpha and Gross Beta
                              Alpha-Emitting Radium Isotopes
                              PART II   CHARACTERISTICS

CHAPTER SEVEN --  INTRODUCTION AND REGULATORY DEFINITIONS

      7.1   Ignitability
      7.2   Corrosivity
      7.3   Reactivity

            Test  Method to Determine Hydrogen Cyanide  Released  from Wastes
            Test  Method to Determine Hydrogen Sulfide  Released  from Wastes

      7.4   Toxicity Characteristic Leaching Procedure
                                    CONTENTS  - 9
                                                                          Revision 2
                                                                      September 1994

-------
CHAPTER EIGHT -- METHODS FOR DETERMINING CHARACTERISTICS

      8.1   Ignitability

            Method 1010:      Pensky-Martens   Closed-Cup   Method   for   Determining
                              Ignitability
            Method 1020A:     Setaflash Closed-Cup Method for Determining Ignitability

      8.2   Corrosivity

            Method 1110:      Corrosivity Toward Steel

      8.3   Reactivity
      8.4   Toxicity

            Method 1310A:     Extraction  Procedure   (EP)  Toxicity  Test  Method  and
                              Structural Integrity Test
            Method 1311:      Toxicity Characteristic Leaching Procedure

APPENDIX -- COMPANY REFERENCES
        NOTE:   A suffix of "A" in the method  number  indicates  revision one
        (the method  has been  revised once).   A suffix of  "B"  in  the method
        number  indicates revision two (the method has been revised twice). In
        order to  properly  document the method  used  for  analysis,  the entire
        method  number Including the suffix letter designation (e.g., A or B)
        must be Identified by the analyst.   A  method  reference found within
        the  RCRA regulations  and the text  of SW-846 methods  and  chapters
        refers  to the latest  promulgated revision of the method, even though
        the method number  does not include the appropriate letter suffix.
                                    CONTENTS -  10                         Revision  2
                                                                      September  1994

-------
                                     VOLUME  TWO
DISCLAIMER
ABSTRACT
TABLE OF CONTENTS
METHOD INDEX AND CONVERSION TABLE
PREFACE
CHAPTER ONE. REPRINTED -- QUALITY CONTROL

      1.0   Introduction
      2.0   QA Project Plan
      3.0   Field Operations
      4.0   Laboratory Operations
      5.0   Definitions
      6.0   References
                                 PART III    SAMPLING

CHAPTER NINE -- SAMPLING PLAN

      9.1   Design and Development
      9.2   Implementation

CHAPTER TEN -- SAMPLING METHODS

            Method 0010:      Modified Method 5 Sampling Train
                  Appendix A:       Preparation of XAD-2 Sorbent Resin
                  Appendix B:       Total Chromatographable Organic Material Analysis
            Method 0020:      Source Assessment Sampling System (SASS)
            Method 0030:      Volatile Organic Sampling Train
                                PART IV   MONITORING

CHAPTER ELEVEN -- GROUND WATER MONITORING

      11.1  Background and Objectives
      11.2  Relationship to the Regulations and to Other Documents
      11.3  Revisions and Additions
      11.4  Acceptable Designs and Practices
      11.5  Unacceptable Designs and Practices


CHAPTER TWELVE -- LAND TREATMENT MONITORING

      12.1  Background
      12.2  Treatment Zone
      12.3  Regulatory Definition
                                    CONTENTS -  11                          Rev is-, or. :
                                                                      September 1994

-------
      12.4  Monitoring and Sampling Strategy
      12.5  Analysis
      12.6  References and Bibliography

CHAPTER THIRTEEN - INCINERATION

      13.1  Introduction
      13.2  Regulatory Definition
      13.3  Waste Characterization Strategy
      13.4  Stack-Gas Effluent Characterization Strategy
      13.5  Additional Effluent Characterization Strategy
      13.6  Selection of Specific Sampling and Analysis Methods
      13.7  References

APPENDIX -- COMPANY REFERENCES
        NOTE;  A  suffix of "A" in the method  number  indicates  revision  one
        (the method has  been  revised once).   A suffix of "B" in  the  method
        number indicates revision two (the method has been revised twice). In
        order to properly  document the method  used  for analysis,  the  entire
        method number including the suffix letter designation (e.g., A or B)
        must be Identified by the analyst.   A  method  reference  found  within
        the  RCRA  regulations  and the text  of SW-846 methods  and chapters
        refers to the latest  promulgated revision of the  method, even  though
        the method number  does not include the appropriate letter suffix.
                                    CONTENTS -  12                         Revision 2
                                                                      September 1994

-------
                      METHOD INDEX AND CONVERSION TABLE
Method Number,
Third Edition
    0010
    0020
    0030
    1010
    1020

    1110
    1310
    1320
    1330
    3005

    3010
    3020
    3040
    3050
    3500

    3510
    3520
    3540
    3550
    3580

    3600
    3610
    3611
    3620
    3630

    3640
    3650
    3660
    3810
    3820

    5030
    5040
    6010
    7000
    7020
Chapter Number,
Third Edition
   Ten
   Ten
   Ten
   Eight (8.1)
   Eight (8.1)

   Eight (8.2)
   Eight (8.4)
   Six
   Six
   Three

   Three
   Three
   Three
   Three
   Four (4.2.1)

   Four (4.2.1)
   Four (4.2.1)
   Four (4.2.1)
   Four (4.2.1)
   Four (4.2.1)

   Four (4.2.2)
   Four (4.2.2)
   Four (4.2.2)
   Four (4.2.2)
   Four (4.2.2)

   Four (4.2.2)
   Four (4.2.2)
   Four (4.2.2)
   Four (4.4)
   Four (4.4)

   Four (4.2.1)
   Four (4.2.1)
   Three
   Three
   Three
Method Number,
Current Revision
Second Edition
0010
0020
0030
1010
1020
1110
1310
1320
1330
3005
3010
3020
3040
3050
None (new method)
3510
3520
3540
3550
None (new method)
None (new method)
None (new method)
3570
None (new method)
None (new method)
None (new method)
None (new method)
None (new method)
5020
None (new method)
5030
3720
6010
7000
7020
Number
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
                               METHOD   INDEX - 1
                                                          Revision       0
                                                          Date   September  1986

-------
                      METHOD INDEX AND CONVERSION TABLE
                                 (Continued)
Method Number,
Third Edition
Chapter Number,
Third Edition
Method Number,
Second Edition
Current Revision
    Number
    7040
    7041
    7060
    7061
    7080

    7090
    7091
    7130
    7131
    7140

    7190
    7191
    7195
    7196
    7197

    7198
    7200
    7201
    7210
    7380

    7420
    7421
    7450
    7460
    7470

    7471
    7480
    7481
    7520
    7550

    7610
    7740
    7741
    7760
    7770
   Three
   Three
   Three
   Three
   Three

   Three
   Three
   Three
   Three
   Three

   Three
   Three
   Three
   Three
   Three

   Three
   Three
   Three
   Three
   Three

   Three
   Three
   Three
   Three
   Three

   Three
   Three
   Three
   Three
   Three

   Three
   Three
   Three
   Three
   Three
  7040
  7041
  7060
  7061
  7080

  7090
  7091
  7130
  7131
  7140

  7190
  7191
  7195
  7196
  7197

  7198
  7200
  7201
  7210
  7380

  7420
  7421
  7450
  7460
  7470

  7471
  7480
  7481
  7520
  7550

  7610
  7740
  7741
  7760
  7770
     0
     0
     0
     0
     0

     0
     0
     0
     0
     0

     0
     0
     0
     0
     0

     0
     0
     0
     0
     0

     0
     0
     0
     0
     0

     0
     0
     0
     0
     0

     0
     0
     0
     0
     0
                               METHOD  INDEX - 2
                                                          Revision      0
                                                          Date  September 1986

-------
                      METHOD INDEX AND CONVERSION TABLE
                                 (Continued)
Method Number,
Third Edition
Chapter Number,
Third Edition
Method Number,
Second Edition
Current Revision
    Number
    7840
    7841
    7870
    7910
    7911

    7950
    8000
    8010
    8015
    8020

    8030
    8040
    8060
    8080
    8090

    8100
    8120
    8140
    8150
    8240

    8250
    8270
    8280
    8310
    9010

    9020
    9022
    9030
    9035
    9036

    9038
    9040
    9041
    9045
    9050
   Three
   Three
   Three
   Three
   Three

   Three
   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)

   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)

   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.1)
   Four  (4.3.2)

   Four  (4.3.2)
   Four  (4.3.2)
   Four  (4.3.2)
   Four  (4.3.3)
   Five

   Five
   Five
   Five
   Five
   Five

   Five
   Six
   Six
   Six
   Six
  7840                   0
  7841                   0
  7870                   0
  7910                   0
  7911                   0

  7950                   0
  None  (new method)      0
  8010                   0
  8015                   0
  8020                   0

  8030                   0
  8040                   0
  8060                   0
  8080                   0
  8090                   0

  8100                   0
  8120                   0
  8140                   0
  8150                   0
  8240                   0

  8250                   0
  8270                   0
  None  (new method)      0
  8310                   0
  9010                   0

  9020                   0
  9022                   0
  9030                   0
  9035                   0
  9036                   0

  9038                   0
  9040                   0
  9041                   0
  9045                   0
  9050                   0
                               METHOD  INDEX - 3
                                                          Revision      0
                                                          Date  September 1986

-------
                      METHOD INDEX AND CONVERSION TABLE
                                 (Continued)
Method Number,
Third Edition
Chapter Number,
Third Edition
Method Number,
Second Edition
Current Revision
    Number
    9060               Five
    9065               Five
   •9066               Five
    9067               Five
    9070               Five

    9071               Five
    9080               Six
    9081               Six
    9090               Six
    9095               Six

    9100               Six
    9131               Five
    9132               Five
    9200               Five
    9250               Five

    9251               Five
    9252               Five
    9310               Six
    9315               Six
    9320               Five

    HCN Test  Method   Seven
        Test  Method   Seven
                       9060
                       9065
                       9066
                       9067
                       9070

                       9071
                       9080
                       9081
                       9090
                       9095

                       9100
                       9131
                       9132
                       9200
                       9250

                       9251
                       9252
                       9310
                       9315
                       9320

                       HCN Test  Method
                           Test  Method
                         0
                         0
                         0
                         0
                         0

                         0
                         0
                         0
                         0
                         0

                         0
                         0
                         0
                         0
                         0

                         0
                         0
                         0
                         0
                         0

                         0
                         0
                               METHOD  INDEX - 4
                                                          Revision      0
                                                          Date  September 1986

-------
   CHAPTER ONE
TABLE OF CONTENTS
Section
1.0
2.0















3.0






















INTRODUCTION 	
QA PROJECT PLAN 	
2.1 DATA QUALITY OBJECTIVES 	
2.2 PROJECT OBJECTIVES 	
2.3 SAMPLE COLLECTION 	
2.4 ANALYSIS AND TESTING 	
2.5 QUALITY CONTROL 	
2.6 PROJECT DOCUMENTATION 	
2.7 ORGANIZATION PERFORMING FIELD OR LABORATORY
OPERATIONS 	
2.7.1 Performance Evaluation 	
2.7.2 Internal Assessment by QA Function 	
2.7.3 External Assessment 	
2.7.4 On-Site Evaluation 	
2.7.4.1 Field Activities 	
2.7.4.2 Laboratory Activities 	
2.7.5 QA Reports 	
FIELD OPERATIONS 	
3.1 FIELD LOGISTICS 	
3.2 EQUIPMENT/ INSTRUMENTATION 	
3.3 OPERATING PROCEDURES 	
3.3.1 Sample Management 	
3.3.2 Reagent/Standard Preparation 	
3.3.3 Decontamination 	
3.3.4 Sample Collection 	
3.3.5 Field Measurements 	
3.3.6 Equipment Calibration And Maintenance . . . .
3.3.7 Corrective Action 	
3.3.8 Data Reduction and Validation 	
3.3.9 Reporting 	
3.3.10 Records Management 	
3.3.11 Waste Disposal 	
3.4 FIELD QA AND QC REQUIREMENTS 	
3.4.1 Control Samples 	
3.4.2 Acceptance Criteria 	
3.4.3 Deviations 	
3.4.4 Corrective Action 	
3.4.5 Data Handling 	
3.5 QUALITY ASSURANCE REVIEW 	
3.6 FIELD RECORDS 	
Page
.... 1
.... 1
.... 2
.... 2
.... 3
.... 3
.... 3
.... 3

.... 4
.... 5
.... 5
.... 5
.... 5
.... 5
.... 6
.... 7
.... 8
.... 8
.... 9
.... 9
.... 9
.... 9
.... 9
.... 10
.... 10
.... 10
.... 10
.... 11
.... 11
.... 11
.... 11
.... 11
.... 11
.... 12
.... 12
.... 12
.... 12
.... 13
.... 13
     ONE -  i                          Revision 1
                                      July 1992

-------
TABLE OF CONTENTS
   (continued)
Section
4.0























5.0
6.0
LABORATORY OPERATIONS 	
4.1 FACILITIES 	
4.2 EQUIPMENT/ INSTRUMENTATION 	
4.3 OPERATING PROCEDURES 	
4.3.1 Sample Management 	
4.3.2 Reagent/Standard Preparation 	
4.3.3 General Laboratory Techniques 	
4.3.4 Test Methods 	
4.3.5 Equipment Calibration and Maintenance . . . .
4.3.6 QC 	
4.3.7 Corrective Action 	
4.3.8 Data Reduction and Validation 	
4.3.9 Reporting 	
4.3.10 Records Management 	
4.3.11 Waste Disposal 	
4.4 LABORATORY QA AND QC PROCEDURES 	
4.4.1 Method Proficiency 	
4.4.2 Control Limits 	
4.4.3 Laboratory Control Procedures 	
4.4.4 Deviations 	
4.4.5 Corrective Action 	
4.4.6 Data Handling 	
4.5 QUALITY ASSURANCE REVIEW 	
4.6 LABORATORY RECORDS 	
DEFINITIONS 	
REFERENCES 	
INDEX 	
Page
.... 14
.... 14
.... 15
.... 15
.... 16
.... 16
.... 16
.... 16
.... 17
.... 17
.... 17
.... 18
.... 18
.... 18
.... 18
.... 18
.... 18
.... 19
.... 19
.... 20
... 20
.... 20
.... 21
.... 21
.... 23
.... 29
.... 30
     ONE  -  ii                          Revision 1
                                       July 1992

-------
                                 CHAPTER ONE
                                QUALITY  CONTROL
1.0  INTRODUCTION
      It  is  the goal  of  the U.S.  Environmental  Protection  Agency's (EPA's)
quality assurance (QA)  program to  ensure that all data be scientifically valid,
defensible,  and of  known precision  and   accuracy.   The  data  should be  of
sufficient known quality to withstand scientific and legal challenge relative to
the use for which the data are obtained.  The QA program is management's tool for
achieving this goal.

      For RCRA analyses,  the recommended minimum requirements for a QA program
and the associated  quality control (QC) procedures  are provided in this chapter.

      The data acquired from QC procedures are used to estimate the quality of
analytical data, to  determine the need for  corrective  action  in  response  to
identified  deficiencies,  and  to interpret  results  after corrective  action
procedures are implemented.   Method-specific QC procedures are incorporated in
the individual methods since they are not  applied  universally.

      A total program to generate data of acceptable quality should include both
a QA component, which encompasses the management procedures and  controls, as well
as an  operational  day-to-day QC component.  This  chapter  defines  fundamental
elements of such a data collection program.  Data  collection efforts involve:

      1.    design of  a  project plan to  achieve  the data  quality  objectives
            (DQOs);

      2.    implementation of the project  plan; and

      3.    assessment of the data to determine if the DQOs are met.

The project plan may  be a sampling and analysis  plan or a waste analysis plan if
it covers  the QA/QC  goals of the Chapter, or  it  may be  a Quality Assurance
Project Plan as described later in this  chapter.

      This chapter identifies the minimal  QC components  that should be used in
the performance  of sampling  and analyses, including  the QC  information  which
should be documented.  Guidance is provided to construct QA programs for field
and laboratory work conducted in support of the RCRA program.


2.0  QA PROJECT PLAN

      It is recommended that  all projects which generate environment-related data
in support  of RCRA  have  a  QA  Project  Plan  (QAPjP)  or  equivalent.   In  some
instances,  a sampling  and   analysis  plan or  a waste  analysis  plan may  be
equivalent if it covers all  of  the  QA/QC  goals outlined in this chapter.   In
addition,  a   separate  QAPjP  need  not  be prepared  for  routine  analyses  or
activities where  the procedures  to  be  followed are  described  in  a  Standard

                                   ONE -  1                          Revision 1
                                                                     July 1992

-------
Operating Procedures manual or similar document  and  include  the elements of a
QAPjP.  These documents should be available and referenced in the documentation
and/or records for the analysis activities.   The term "QAPjP"  in this chapter
refers to any of these QA/QC documents.

      The QAPjP should detail the QA/QC goals  and protocols for a specific data
collection activity.   The  QAPjP sets forth  a plan for sampling  and analysis
activities that will generate data of a quality commensurate with their intended
use.   QAPjP  elements  should include  a description  of the  project  and  its
objectives; a statement of  the DQOs of the project; identification of those in-
volved  in the  data collection  and  their  responsibilities  and  authorities;
reference  to  (or inclusion  of)  the  specific  sample collection  and analysis
procedures that will  be followed for  all aspects  of the project; enumeration of
QC procedures  to  be  followed;  and descriptions  of all  project documentation.
Additional elements should  be included  in  the QAPjP  if needed  to address all
quality related  aspects  of the  data collection  project.   Elements  should be
omitted only when  they are inappropriate  for the project or when  absence of those
elements  will  not affect the quality  of data obtained  for the  project (see
reference 1).

      The role and  importance of DQOs and  project documentation are discussed
below in Sections 2.1  through 2.6.  Management and organization play a critical
role in determining the effectiveness of a QA/QC program and ensuring that all
required  procedures are  followed.   Section 2.7  discusses the   elements of an
organization's QA program that  have been found to ensure an effective program.
Field operations and laboratory operations (along  with  applicable QC procedures)
are discussed  in Sections 3 and  4,  respectively.


2.1  DATA QUALITY OBJECTIVES

      Data quality objectives (DQOs)  for the data collection activity describe
the overall level of uncertainty that a decision-maker is  willing to accept in
results derived from environmental data.  This  uncertainty is used to specify the
quality of  the measurement  data  required,  usually in terms  of objectives for
precision, bias, representativeness,  comparability and completeness.  The DQOs
should be defined prior to the initiation of the field and laboratory work.  The
field and  laboratory organizations performing the  work  should  be aware of the
DQOs so that their personnel may make  informed decisions during the course of the
project to attain those  DQOs.  More  detailed  information  on  DQOs is available
from the U.S.  EPA Quality Assurance  Management  Staff (QAMS)  (see  references 2 and
4).


2.2  PROJECT OBJECTIVES

      A statement of  the project  objectives and how  the objectives  are to be
attained  should be  concisely stated  and  sufficiently detailed  to permit clear
understanding  by  all  parties involved  in the data  collection effort.   This
includes a statement of what problem is to be solved and the information required


                                   ONE  -  2                          Revision 1
                                                                     July 1992

-------
in the process.  It also includes appropriate statements of the DQOs (i.e., the
acceptable level of uncertainty  in the information).


2.3  SAMPLE COLLECTION

      Sampling  procedures,  locations,  equipment, and  sample  preservation and
handling requirements  should be  specified  in the QAPjP.   Further  details on
quality assurance procedures for field  operations are described in Section 3 of
this chapter.  The OSW  is developing policies and procedures for sampling in a
planned  revision  of Chapter Nine of  this  manual.   Specific procedures for
groundwater sampling are provided in Chapter  Eleven of this manual.


2.4  ANALYSIS AND TESTING

      Analytes and properties of  concern, analytical and testing procedures to
be employed,  required detection limits,  and requirements for precision and bias
should be specified.  All applicable regulatory requirements and the project DQOs
should be considered when developing the specifications. Further details on the
procedures for analytical operations are described in Section  4 of this chapter.


2.5  QUALITY CONTROL

      The quality  assurance program should  address both  field and  laboratory
activities.  Quality control procedures should be specified for estimating the
precision and bias of the data.  Recommended minimum requirements for QC samples
have been established by EPA and  should be met in order to satisfy recommended
minimum criteria for acceptable data quality.  Further details  on procedures for
field and laboratory operations are described  in Sections 3 and 4, respectively,
of this chapter.


2.6  PROJECT DOCUMENTATION

      Documents should  be prepared and maintained in conjunction with the data
collection effort.  Project documentation should be sufficient to allow review
of all aspects of the work being  performed.   The QAPjP discussed in  Sections 3
and 4 is one important  document  that should  be maintained.

      The  length  of storage  time  for project  records   should  comply  with
regulatory  requirements,  organizational  policy,  or  project  requirements,
whichever is more stringent.  It  is recommended that documentation be stored for
three years from submission of the project final report.

      Documentation  should  be  secured   in  a   facility   that   adequately
addresses/minimizes  its deterioration for  the length of time  that it is to be
retained.  A system allowing for  the expedient retrieval  of information should
exist.
                                    ONE - 3                         Revision 1
                                                                     July 1992

-------
      Access  to  archived  information should  be controlled  to maintain  the
integrity  of  the  data.    Procedures should  be developed  to  identify  those
individuals with access to the data.
2.7  ORGANIZATION PERFORMING FIELD OR LABORATORY OPERATIONS

      Proper design and structure  of the organization facilitates effective and
efficient transfer of information and helps to prevent important procedures from
being overlooked.

      The  organizational  structure,  functional  responsibilities,   levels  of
authority,  job  descriptions,  and  lines  of  communication  for  all  project
activities should be established and documented.  One person may cover more than
one  organizational  function.   Each project  participant  should  have  a  clear
understanding of his or her  duties and responsibilities and the relationship of
those responsibilities to the overall data collection effort.

      The management of each organization participating in a project involving
data collection activities should  establish  that organization's operational and
QA policies. This information should be documented in the QAPjP.  The management
should ensure that (1)  the appropriate methodologies are followed as documented
in   the   QAPjPs;   (2)   personnel   clearly   understand   their  duties   and
responsibilities;  (3)  each  staff  member  has  access  to  appropriate  project
documents;  (4)  any  deviations  from the QAPjP  are communicated to  the project
management  and documented;  and  (5) communication  occurs between  the field,
laboratory, and project management,  as specified in the QAPjP.  In addition, each
organization should ensure  that  their activities do not  increase  the risk to
humans or the environment  at or about the project location.  Certain projects may
require specific policies  or a Health and Safety Plan to provide this  assurance.

      The management of the participating field or laboratory organization should
establish personnel qualifications  and training requirements for the project.
Each person  participating in the  project  should have the education, training,
technical knowledge, and  experience,  or  a combination  thereof,  to  enable that
individual to perform assigned  functions.   Training  should be provided  for each
staff  member  as  necessary  to perform  their  functions   properly.    Personnel
qualifications  should  be documented in  terms of education, experience,  and
training,   and   periodically   reviewed   to   ensure   adequacy   to  current
responsibilities.

      Each  participating  field organization or laboratory organization should
have a  designated QA  function (i.e., a  team  or individual  trained  in QA) to
monitor  operations  to  ensure  that the  equipment,  personnel,  activities,
procedures,  and  documentation conform with  the QAPjP.  To the extent possible,
the QA monitoring function should be entirely separate from, and independent of,
personnel  engaged in  the work being  monitored.    The  QA function  should be
responsible  for  the QA review.
                                    ONE - 4                         Revision 1
                                                                     July  1992

-------
      2.7.1  Performance Evaluation

      Performance evaluation studies are used  to measure the performance of the
laboratory on  unknown samples.  Performance  evaluation  samples are typically
submitted to the laboratory as blind samples by an independent outside source.
The  results  are  compared to  predetermined  acceptance  limits.   Performance
evaluation samples  can  also  be submitted to the laboratory as  part of the QA
function during internal assessment of laboratory performance.  Records of all
performance evaluation studies should be maintained by the  laboratory.  Problems
identified through  participation in performance evaluation studies should be
immediately investigated and corrected.

      2.7.2  Internal Assessment by QA Function

      Personnel performing field and laboratory activities are responsible for
continually monitoring  individual  compliance  with  the QAPjP.   The QA function
should review procedures, results and calculations to determine compliance with
the  QAPjP.   The  results of  this  internal assessment  should be  reported to
management with requirements for a plan to correct observed deficiencies.

      2.7.3  External Assessment

      The field and laboratory activities may  be reviewed  by personnel external
to the  organization.   Such an assessment  is  an extremely valuable method for
identifying overlooked problems.  The results  of the external assessment should
be submitted  to management with  requirements  for  a plan  to  correct observed
deficiencies.

      2.7.4  On-Site  Evaluation

      On-site evaluations may be conducted  as part of both  internal  and external
assessments.   The  focus  of  an  on-site  evaluation  is  to evaluate the degree of
conformance of project activities with the applicable QAPjP. On-site evaluations
may  include, but  are not limited to,  a  complete review  of facilities,  staff,
training, instrumentation, procedures,  methods, sample collection,  analyses, QA
policies and procedures  related to the generation of environmental data.  Records
of each evaluation should include the date of the evaluation, location, the areas
reviewed,  the  person performing the  evaluation,  findings and  problems,  and
actions recommended and taken  to  resolve problems.   Any problems  identified that
are  likely to  affect  data  integrity   should  be  brought immediately to  the
attention of management.

            2.7.4.1   Field Activities

      The review of field activities should be conducted by one or  more persons
knowledgeable  in  the activities being  reviewed  and include  evaluating,  at  a
minimum, the following subjects:

      Completeness  of  Field  Reports   --  This review  determines  whether  all
      requirements  for field activities in the QAPjP have  been fulfilled, that
      complete  records  exist  for each  field  activity,  and that the procedures

                                    ONE -  5                         Revision  1
                                                                     July 1992

-------
      specified  in  the  QAPjP  have  been  implemented.    Emphasis  on  field
      documentation  will  help assure sample integrity and sufficient technical
      information  to   recreate  each  field  event.     The  results  of  this
      completeness check should be documented, and environmental  data affected
      by incomplete  records should be identified.

      Identification of Valid Samples -- This review involves interpretation and
      evaluation  of the field records  to  detect problems  affecting  the repre-
      sentativeness  of environmental  samples.  Examples  of items  that  might
      indicate potentially invalid  samples  include improper  well  development,
      improperly  screened wells,  instability of pH or conductivity, and collec-
      tion of volatiles  near  internal  combustion  engines.   The  field records
      should be evaluated against the QAPjP and SOPs.  The reviewer should docu-
      ment the sample validity and  identify the environmental  data  associated
      with any poor  or incorrect field work.

      Correlation of  Field Test  Data  -- This  review  involves  comparing  any
      available results of  field  measurements obtained by more than one method.
      For example,  surface geophysical methods should  correlate with  direct
      methods  of site  geologic  characterization  such  as  lithologic  logs
      constructed during drilling operations.

      Identification of Anomalous Field Test Data  -- This review identifies any
      anomalous field  test  data.   For example, a water temperature for one well
      that is  5  degrees higher  than  any other well  temperature in  the  same
      aquifer should  be  noted.   The  reviewer should evaluate the  impact  of
      anomalous field  measurement results  on the associated environmental data.

      Validation  of Field  Analyses  --  This  review validates  and  documents all
      data from  field analysis that  are  generated  in situ  or from  a  mobile
      laboratory  as  specified in  Section 2.7.4.2.  The reviewer should document
      whether the QC checks meet the acceptance criteria, and whether corrective
      actions were  taken for any analysis  performed when  acceptance criteria
      were exceeded.

            2.7.4.2   Laboratory Activities

      The review  of laboratory data should be conducted  by one or more persons
knowledgeable in  laboratory activities and include  evaluating,  at a minimum, the
following subjects:

      Completeness of Laboratory Records --  This review determines whether: (1)
      all samples and  analyses required by  the QAPjP have been processed,  (2)
      complete records exist for  each  analysis and the  associated QC samples,
      and that (3) the procedures specified  in the QAPjP have been implemented.
      The  results  of  the  completeness  check  should  be  documented,   and
      environmental  data affected by incomplete records  should be identified.

      Evaluation  of Data with  Respect  to  Detection and  Quantitation Limits  --
      This review compares analytical results to required quantitation limits.
      Reviewers should document instances where detection or quantitation limits

                                    ONE -  6                          Revision 1
                                                                     July 1992

-------
      exceed  regulatory  limits,  action  levels,   or  target  concentrations
      specified in the QAPjP.

      Evaluation of Data with Respect to Control Limits -- This review compares
      the   results  of QC and  calibration  check samples to  control  criteria.
      Corrective  action should  be  implemented  for data  not within  control
      limits. The reviewer should check that corrective action reports, and the
      results  of reanalysis,  are   available.    The  review  should  determine
      whether samples associated with  out-of-control QC data are identified in
      a written  record  of the data review, and  whether an  assessment  of the
      utility of such analytical results  is recorded.

      Review of Holding Time Data -- This review compares sample holding times
      to those required by the QAPjP,  and notes all  deviations.

      Review of Performance Evaluation (PE) Results  -- PE study results can be
      helpful in evaluating the impact of  out-of-control conditions. This review
      documents  any  recurring trends or  problems  evident  in  PE  studies  and
      evaluates their effect on environmental  data.

      Correlation of  Laboratory Data  --  This review  determines  whether  the
      results of data obtained  from related laboratory  tests,  e.g.,  Purgeable
      Organic Hal ides (POX) and Volatile  Organics,  are documented,  and whether
      the significance of any differences is discussed in the reports.

      2.7.5  QA Reports

      There should be periodic reporting  of pertinent  QA/QC information to the
project management to allow  assessment of  the  overall effectiveness  of the QA
program.  There are three major types  of  QA reports  to project management:

      Periodic Report on Key  QA Activities -- Provides  summary of key QA activi-
      ties during the  period, stressing measures that are  being taken to improve
      data  quality;  describes  significant  quality   problems   observed  and
      corrective actions  taken;  reports  information regarding any changes in
      certification/accreditation status; describes  involvement in resolution of
      quality issues  with  clients  or  agencies; reports  any  QA  organizational
      changes;  and  provides  notice of the distribution of  revised  documents
      controlled by the QA organization (i.e.,  procedures).

      Report on Measurement  Quality  Indicators  -- Includes the assessment of QC
      data gathered over the period,  the  frequency of  analyses  repeated due to
      unacceptable QC performance,  and, if possible, the reason  for  the unac-
      ceptable performance and corrective action  taken.

      Reports on QA Assessments --  Includes the results  of  the assessments and
      the plan  for  correcting identified deficiencies;  submitted  immediately
      following any internal  or external  on-site  evaluation  or upon receipt of
      the results of any performance evaluation studies.
                                   ONE - 7                          Revision  1
                                                                     July  1992

-------
3.0  FIELD OPERATIONS

      The field  operations  should be  conducted  in such  a  way as  to provide
reliable information  that  meets the DQOs.   To achieve this,  certain minimal
policies and procedures  should be implemented.  The  OSW  is considering revisions
of Chapter Nine and Eleven of this manual.  Supplemental  information and guidance
is available in  the  RCRA Ground-Water Monitoring Technical Enforcement Guidance
Document (TEGD)  (Reference  3).   The project documentation  should  contain the
information specified below.


3.1  FIELD LOGISTICS

      The QAPjP should describe  the type(s) of  field operations to be performed
and the  appropriate  area(s) in  which  to  perform  the  work.   The  QAPjP should
address ventilation,  protection  from extreme weather and temperatures, access to
stable power, and provision for water and gases of required  purity.

      Whenever practical, the sampling  site facilities  should be examined prior
to the start of work to ensure that all required items are  available.  The actual
area of sampling should  be examined to ensure that trucks, drilling equipment,
and personnel have adequate access to the site.

      The determination  as to whether sample shipping is necessary should be made
during planning  for  the project.   This need is established by evaluating the
analyses to be performed,  sample holding times,  and location  of  the site and the
laboratory.  Shipping or transporting of samples to a laboratory should be done
within a timeframe such that recommended holding times are met.

      Samples should be  packaged, labelled, preserved (e.g.,  preservative added,
iced,  etc.),  and documented  in an area  which is  free  of  contamination and
provides for secure  storage. The level  of custody and whether sample storage is
needed should be addressed  in the QAPjP.

      Storage areas for solvents, reagents,  standards,  and reference materials
should  be  adequate  to  preserve  their  identity,  concentration,  purity,  and
stability prior to use.

      Decontamination of  sampling  equipment  may be performed  at  the location
where  sampling occurs,  prior to going  to the  sampling site, or in  designated
areas near the sampling  site.  Project documentation should specify where and how
this work is accomplished.  If decontamination  is to be done at the site, water
and  solvents of appropriate  purity  should  be  available.    The  method  of
accomplishing decontamination,  including the required  materials, solvents, and
water purity  should be  specified.

      During  the sampling process and during  on-site or in situ analyses, waste
materials are sometimes  generated.  The method for storage and disposal of these
waste  materials  that  complies  with   applicable   local,   state  and  Federal
regulations should be specified. Adequate facilities should  be  provided for the
collection and storage of all wastes, and these  facilities should be operated so

                                    ONE -  8                          Revision 1
                                                                     July 1992

-------
as  to  minimize environmental  contamination.    Waste  storage  and  disposal
facilities should comply with applicable federal,  state,  and local regulations.

      The location of long-term and short-term storage for field records,  and the
measures to ensure the integrity of the data should be specified.


3.2  EQUIPMENT/INSTRUMENTATION

      The equipment, instrumentation, and supplies at the sampling site should
be specified and should be appropriate to accomplish the activities planned.  The
equipment and  instrumentation should  meet  the requirements of  specifications,
methods, and procedures as specified in the QAPjP.


3.3  OPERATING PROCEDURES

      The QAPjP should describe or make reference to all field  activities that
may affect data quality.  For routinely performed activities, standard operating
procedures (SOPs) are  often prepared to ensure  consistency and to  save time and
effort in preparing QAPjPs.  Any deviation from an established procedure during
a  data  collection activity  should  be  documented.   The procedures  should  be
available for the  indicated  activities,  and  should  include,  at a minimum,  the
information described below.

      3.3.1  Sample Management

      The numbering and labeling system,  chain-of-custody procedures, and  how the
samples  are  to  be tracked  from collection  to  shipment  or  receipt  by  the
laboratory should be specified. Sample management procedures should also specify
the  holding  times, volumes  of sample  required  by  the laboratory,  required
preservatives, and shipping requirements.

      3.3.2  Reagent/Standard Preparation

      The procedures describing  how  to  prepare  standards  and reagents should be
specified.  Information concerning specific grades of materials  used in reagent
and standard preparation, appropriate glassware and containers  for preparation
and storage,  and labeling and record keeping for stocks and dilutions should be
included.

      3.3.3  Decontamination

      The procedures describing decontamination of  field equipment  before  and
during the  sample  collection process  should  be  specified.   These  procedures
should include cleaning materials used, the order of washing  and rinsing with the
cleaning materials, requirements for protecting or covering cleaned equipment,
and procedures for disposing of cleaning materials.
                                   ONE  - 9                          Revision 1
                                                                     July 1992

-------
      3.3.4  Sample Collection

      The  procedures  describing  how  the  sampling  operations   are  actually
performed  in  the  field  should be specified.   A simple  reference to standard
methods is not sufficient,  unless  a procedure is performed exactly as described
in the published method.   Methods from  source  documents  published by the EPA,
American Society for  Testing  and  Materials, U.S. Department  of  the Interior,
National  Water  Well  Association,  American   Petroleum  Institute,  or  other
recognized organizations with appropriate expertise should be used, if possible.
The procedures for sample collection should include at least the following:

   •  Applicability of the procedure,

      Equipment required,

   •  Detailed  description of  procedures   to  be followed  in collecting  the
      samples,

   •  Common problems encountered and corrective actions to be followed, and

   •  Precautions to be taken.

      3.3.5  Field Measurements

      The  procedures  describing all methods used in  the field to determine a
chemical or physical  parameter should be described  in detail.   The procedures
should address criteria from Section 4,  as appropriate.

      3.3.6  Equipment Calibration And Maintenance

      The  procedures  describing  how  to   ensure   that  field  equipment  and
instrumentation  are  in working  order  should  be specified.   These  describe
calibration procedures  and schedules,  maintenance  procedures and  schedules,
maintenance logs,  and  service  arrangements for  equipment.   Calibration  and
maintenance of field equipment and instrumentation should be in accordance with
manufacturers' specifications or  applicable test  specifications  and should be
documented.

      3.3.7  Corrective Action

      The procedures describing how  to identify and  correct  deficiencies in the
sample collection process  should  be specified.   These should  include specific
steps  to  take  in  correcting  deficiencies   such  as  performing  additional
decontamination  of equipment,  resampling, or additional  training of field
personnel.  The procedures should  specify that  each  corrective action should be
documented with a description of the deficiency and the corrective  action taken,
and should include  the person(s)  responsible  for  implementing the corrective
action.
                                   ONE - 10                         Revision 1   i
                                                                     July 1992   1

-------
      3.3.8  Data Reduction and  Validation

      The procedures describing  how  to  compute results from field measurements
and to review and validate these data should be  specified.  They  should include
all formulas  used to  calculate  results and procedures  used  to independently
verify that field measurement results are correct.

      3.3.9  Reporting

      The procedures describing  the  process for reporting the results of field
activities should be specified.

      3.3.10 Records Management

      The  procedures describing the means  for generating,  controlling,  and
archiving  project-specific records  and  field  operations  records  should  be
specified.  These procedures should detail record generation and control and the
requirements for record retention,  including type, time,  security, and retrieval
and disposal authorities.

      Project-specific  records  relate  to field work  performed  for a project.
      These records may include correspondence,  chain-of-custody  records, field
      notes, all reports issued  as a result of the work, and procedures used.

      Field operations records document  overall field operations  and may include
      equipment performance and maintenance logs, personnel  files, general field
      procedures, and corrective action reports.

      3.3.11 Waste Disposal

      The  procedures  describing the methods  for disposal of  waste materials
resulting from field operations  should  be specified.


3.4  FIELD QA AND QC REQUIREMENTS

      The  QAPjP  should describe how the  following   elements of the  field  QC
program will be implemented.

      3.4.1  Control Samples

      Control  samples  are QC  samples  that are  introduced  into a  process  to
monitor the performance of the system. Control samples, which may  include blanks
(e.g.,  trip,  equipment,  and   laboratory),  duplicates,  spikes,   analytical
standards, and reference materials, can  be used  in different phases of the data
collection process beginning with sampling and continuing through transportation,
storage, and analysis.

      Each day  of sampling,  at  least  one  field  duplicate and  one  equipment
rinsate should be collected for  each matrix sampled.  If this frequency is not
appropriate for the sampling  equipment and method, then the appropriate changes

                                   ONE  - 11                         Revision 1
                                                                     July 1992

-------
should be  cTearly identified in  the QAPjP.   When samples are  collected for
volatile organic analysis, a trip blank  is  also recommended for each day that
samples are collected.   In addition,  for  each  sampling batch (20 samples of one
matrix type), enough volume should be collected for at least one sample so as to
allow the laboratory to prepare one matrix spike and either one matrix duplicate
or one matrix spike duplicate for each analytical method employed.  This means
that the following control samples are recommended:

      •Field duplicate (one per day per matrix type)
      •Equipment rinsate  (one per day per matrix type)
      •Trip blank (one per day,  volatile organics only)
      •Matrix spike (one per batch [20 samples of each matrix type])
      •Matrix duplicate or matrix spike duplicate (one per batch)

Additional control samples may be necessary in order to assure data quality to
meet the project-specific DQOs.   '

      3.4.2  Acceptance Criteria

      Procedures  should  be in  place for establishing  acceptance criteria for
field activities described in the QAPjP.  Acceptance criteria may  be qualitative
or  quantitative.    Field  events  or  data  that fall  outside of established
acceptance criteria may indicate a problem with the sampling process that should
be investigated.

      3.4.3  Deviations

      All deviations from plan  should be documented as to the extent of, and
reason  for,  the  deviation.   Any  activity not  performed  in  accordance with
procedures or QAPjPs is considered a deviation from plan.   Deviations from plan
may or may not affect data quality.

      3.4.4  Corrective Action

      Errors, deficiencies, deviations, certain field events,  or  data that fall
outside  established  acceptance  criteria should be  investigated.   In some in-
stances,  corrective  action may be  needed  to resolve the  problem and restore
proper  functioning  to  the system.   The investigation of  the  problem and any
subsequent corrective action taken should be  documented.

      3.4.5  Data Handling

      All  field  measurement  data  should  be  reduced  according  to  protocols
described or referenced in the QAPjP.  Computer programs used for  data reduction
should be validated before use and verified on a regular basis.  All information
used in the calculations  should be recorded to  enable reconstruction of the final
result  at a  later date.

      Data should be reported in accordance with the requirements of the end-user
as described  in the  QAPjP.


                                   ONE  -  12                         Revision  1
                                                                     July 1992

-------
3.5  QUALITY ASSURANCE REVIEW

      The QA Review consists of internal and external  assessments  to ensure that
QA/QC procedures  are in use and  to  ensure that field  staff  conform to these
procedures.  QA review should be  conducted as deemed  appropriate  and necessary.


3.6  FIELD RECORDS

      Records provide the direct evidence and support for the necessary technical
interpretations, judgments, and discussions concerning project activities. These
records, particularly those that  are  anticipated to be used as  evidentiary data,
should directly support current or ongoing technical  studies and  activities and
should provide  the  historical  evidence needed  for later reviews and analyses.
Records should  be legible,  identifiable,  and retrievable and  protected against
damage, deterioration, or  loss.  The discussion in this section  (3.6) outlines
recommended procedures  for record  keeping.   Organizations which conduct field
sampling  should develop appropriate  record keeping procedures  which  satisfy
relevant technical  and legal requirements.

      Field records generally consist of bound  field  notebooks with prenumbered
pages,  sample  collection  forms,  personnel  qualification  and training forms,
sample location maps, equipment maintenance and  calibration  forms,  chain-of-
custody forms,  sample analysis  request forms,  and field change request forms.
All records should   be written in  indelible ink.

      Procedures for  reviewing, approving, and  revising field  records should be
clearly defined, with the  lines of authority included.  It is recommended that
all documentation errors should be corrected by  drawing a single line through the
error  so  it  remains legible  and  should  be  initialed  by  the  responsible
individual, along  with  the date of change.  The  correction  should  be written
adjacent to the error.

      Records should  include (but  are  not limited to) the following:

      Calibration Records & Traceability of Standards/Reagents  --  Calibration is
      a reproducible reference  point to  which  all  sample measurements can be
      correlated.   A sound calibration program should  include  provisions for
      documentation of frequency, conditions, standards, and records reflecting
      the  calibration history  of  a  measurement  system.  The accuracy of the
      calibration standards  is important  because all data will be in reference
      to  the  standards  used.    A  program  for  verifying  and documenting the
      accuracy  of all working standards against primary grade standards should
      be routinely  followed.

      Sample Collection  -- To ensure maximum utility  of the sampling effort and
      resulting data, documentation  of the sampling  protocol, as performed in
      the field, is essential.  It  is recommended that sample collection records
      contain,  at a  minimum,  the names  of persons conducting  the  activity,
      sample  number, sample  location, equipment  used,  climatic  conditions,
      documentation  of  adherence to protocol,  and  unusual observations.   The

                                   ONE -  13                         Revision 1
                                                                     July 1992

-------
      actual  sample collection record is usually one of the following:  a bound
      field notebook with prenumbered  pages,  a pre-printed  form,  or digitized
      information on a computer tape or disc.

      Chain-of-Custody Records --  The chain-of-custody involving the possession
      of samples from  the time they are  obtained  until  they are  disposed  or
      shipped off-site should  be documented as specified  in the QAPjP and should
      include the following information:  (1) the  project name;  (2) signatures
      of samplers;  (3)  the sample  number,  date and time of collection, and grab
      or composite  sample  designation;  (4) signatures of individuals involved in
      sample  transfer; and  (5)  if applicable,  the air bill  or  other shipping
      number.

      Maps and Drawings -- Project planning documents and reports often contain
      maps.  The maps  are used to document  the location of  sample collection
      points  and monitoring wells and as a  means of  presenting environmental
      data.  Information used  to prepare maps and drawings is normally obtained
      through  field  surveys,  property  surveys, surveys  of  monitoring  wells,
      aerial  photography or photogrammetric mapping.   The final, approved maps
      and/or  drawings should have  a revision  number and date and should be sub-
      ject to the same controls as other project  records.

      QC Samples -- Documentation  for generation of QC  samples, such as trip and
      equipment rinsate blanks, duplicate samples,  and any field spikes should
      be maintained.

      Deviations --  All  deviations  from procedural   documents  and  the  QAPjP
      should  be recorded  in the site logbook.

      Reports  -- A  copy of  any  report  issued and  any  supporting documentation
      should  be retained.
4.0  LABORATORY OPERATIONS

      The laboratory should conduct its operations in such a way as to provide
reliable information.  To achieve this,  certain minimal policies and procedures
should be implemented.


4.1  FACILITIES

      The QAPjP  should  address  all  facility-related  issues  that  may  impact
project  data  quality.    Each   laboratory  should  be of  suitable  size  and
construction to facilitate the proper conduct of the analyses.  Adequate bench
space or working  area per analyst should be provided.   The  space requirement per
analyst depends on the equipment  or  apparatus that is  being utilized, the number
of samples that the analyst is expected to handle at any one time,  and the number
of operations that are to be performed concurrently by a single analyst.  Other
issues to be considered include,  but are not limited to, ventilation, lighting,


                                    ONE  -  14                         Revision 1
                                                                     July 1992

-------
control  of dust and drafts,  protection from extreme temperatures, and access to
a source of stable power.

      Laboratories should be designed so that there  is  adequate separation of
functions to ensure  that no laboratory  activity has an  adverse  effect on the
analyses. The laboratory may require specialized facilities such as a perchloric
acid hood or glovebox.

      Separate space for laboratory operations and appropriate ancillary support
should be provided,  as  needed,  for the  performance  of  routine  and specialized
procedures.

      As  necessary  to  ensure  secure  storage  and   prevent  contamination  or
misidentification, there should be adequate  facilities  for receipt and storage
of samples.   The level   of  custody required  and any special requirements for
storage such as refrigeration should  be  described  in planning documents.

      Storage areas for reagents, solvents,  standards,  and reference materials
should  be  adequate  to   preserve  their  identity,  concentration, purity,  and
stability.

      Adequate facilities should be provided for the collection and storage of
all  wastes,  and these facilities should  be operated  so  as to minimize environ-
mental contamination. Waste storage and disposal facilities should comply with
applicable federal, state, and local  regulations.

      The location of long-term and short-term storage of  laboratory records and
the measures to ensure the integrity  of  the  data should be specified.


4.2  EQUIPMENT/INSTRUMENTATION

      Equipment and instrumentation should meet the requirements and specifica-
tions of  the  specific test methods and other  procedures as specified  in the
QAPjP.  The  laboratory should maintain an equipment/instrument description list
that includes  the manufacturer, model  number,  year of purchase, accessories, and
any modifications, updates, or upgrades  that have  been  made.


4.3  OPERATING PROCEDURES

      The QAPjP should describe or make  reference  to all  laboratory activities
that may affect data quality. For routinely performed activities,  SOPs are often
prepared to ensure consistency  and to save time and effort in preparing QAPjPs.
Any deviation from an established procedure  during  a data collection activity
should be documented.  It is recommended that procedures  be available for the
indicated activities, and  include,  at  a minimum,  the  information  described
below.
                                   ONE - 15                         Revision 1
                                                                     July 1992

-------
      4.3.1  Sample Management

      The procedures describing the receipt,  handling,  scheduling,  and storage
of samples should be specified.

      Sample Receipt and Handling -- These procedures describe the  precautions
      to be used in opening sample  shipment containers  and  how to  verify that
      chain-of-custody has been maintained, examine  samples  for damage,  check
      for  proper preservatives  and  temperature,  and  log  samples   into  the
      laboratory sample streams.

      Sample Scheduling --  These procedures  describe the sample scheduling in
      the laboratory and  includes procedures used to ensure that holding time
      requirements are met.

      Sample Storage -- These procedures describe the  storage  conditions for all
      samples, verification and documentation of daily storage temperature, and
      how to  ensure that  custody  of  the samples is maintained while  in  the
      laboratory.

      4.3.2  Reagent/Standard Preparation

      The procedures describing how to prepare standards and reagents should be
specified.  Information concerning  specific grades of materials used in reagent
and standard preparation, appropriate glassware  and  containers for  preparation
and storage, and labeling and recordkeeping for  stocks and dilutions should be
included.

      4.3.3  General Laboratory Technigues

      The procedures describing all essentials of laboratory operations that are
not addressed elsewhere should be specified.   These  techniques should include,
but are not limited to, glassware cleaning procedures, operation of analytical
balances, pipetting techniques, and use of volumetric glassware.

      4.3.4  Test Methods

      Procedures  for test methods  describing  how the  analyses are  actually
performed in the laboratory should be specified.   A simple  reference to standard
methods is not sufficient, unless the analysis is  performed exactly as described
in the  published method.  Whenever methods  from SW-846  are not appropriate,
recognized methods  from source documents published by the EPA, American Public
Health Association (APHA), American Society for Testing and Materials (ASTM), the
National  Institute  for  Occupational  Safety  and Health  (NIOSH),  or  other
recognized organizations with  appropriate expertise should be used, if possible.
The documentation  of the actual laboratory procedures  for  analytical methods
should include the  following:

      Sample  Preparation  and  Analysis Procedures --  These  include  applicable
      holding time, extraction, digestion, or preparation steps as  appropriate
      to  the  method; procedures for  determining the appropriate  dilution to

                                   ONE - 16                         Revision  1
                                                                     July 1992

-------
      analyze;  and  any  other  information  required to  perform the  analysis
      accurately and consistently.

      Instrument Standardization -- This includes concentration(s) and frequency
      of analysis of  calibration  standards,  linear range of the  method,  and
      calibration acceptance criteria.

      Sample Data -- This includes recording requirements and documentation in-
      cluding sample identification number,  analyst, data verification, date of
      analysis and verification,  and  computational method(s).

      Precision and Bias -- This  includes all  analytes for which the method is
      applicable and the conditions  for use of this information.

      Detection  and Reporting  Limits  -- This  includes  all  analytes  in  the
      method.

      Test-Specific QC   --  This   describes  QC  activities  applicable to  the
      specific test and references any  applicable QC procedures.

      4.3.5  Equipment Calibration and  Maintenance

      The procedures  describing  how  to  ensure  that laboratory  equipment  and
instrumentation  are in  working order  should  be specified.   These procedures
include  calibration  procedures   and   schedules,  maintenance  procedures  and
schedules,  maintenance logs, service  arrangements for all  equipment, and spare
parts available in-house.  Calibration  and maintenance of laboratory equipment
and instrumentation should be in  accordance with manufacturers'  specifications
or applicable test specifications  and should be documented.

      4.3.6  QC

      The type,  purpose,  and frequency  of  QC  samples  to  be analyzed  in  the
laboratory and the acceptance criteria  should  be specified.  Information should
include  the  applicability of  the QC  sample  to  the  analytical process,  the
statistical  treatment  of the data,  and the responsibility of laboratory staff and
management in generating and using the  data.   Further details on development of
project-specific QC protocols are  described in Section 4.4.

      4.3.7  Corrective Action

      The procedures describing how to  identify  and correct deficiencies in  the
analytical  process should be specified.  These should include specific steps to
take in  correcting  the  deficiencies  such as preparation of  new standards  and
reagents,  recalibration  and  restandardization  of equipment,  reanalysis  of
samples,  or  additional   training of  laboratory  personnel  in  methods  and
procedures.   The procedures should specify that each  corrective action should be
documented with a description of the deficiency and the corrective action taken,
and should  include  the  person(s)  responsible  for  implementing  the corrective
action.
                                   ONE -  17                         Revision 1
                                                                     July 1992

-------
      4.3.8  Data Reduction and Validation

      The procedures describing how to review and  validate  the data should be
specified.  They should  include procedures for  computing and interpreting the
results from QC samples, and independent procedures to verify that the analytical
results are reported correctly.  In addition,  routine procedures used to monitor
precision and bias,  including evaluations of reagent, equipment rinsate, and trip
blanks,  calibration standards, control  samples,  duplicate and matrix  spike
samples, and surrogate  recovery,  should  be  detailed in  the procedures.   More
detailed validation procedures  should be performed when  required in the contract
or QAPjP.

      4.3.9  Reporting

      The procedures describing the process for reporting  the analytical results
should be specified.

      4.3.10 Records Management

      The  procedures describing  the  means   for  generating,  controlling,  and
archiving laboratory records should be specified.  The procedures should detail
record generation and control, and  the requirements for record retention, includ-
ing type, time, security, and retrieval  and  disposal authorities.

      Project-specific  records may  include  correspondence,  chain-of-custody
      records,  request for analysis, calibration  data records, raw and finished
      analytical and QC data, data reports,  and  procedures used.

      Laboratory operations records may include laboratory notebooks, instrument
      performance logs and maintenance logs in bound  notebooks with prenumbered
      pages; laboratory  benchsheets;  software documentation;  control  charts;
      reference material certification; personnel files;  laboratory procedures;
      and corrective action reports.

      4.3.11 Waste Disposal

      The procedures describing the methods for disposal  of chemicals including
standard and reagent solutions, process waste, and  samples should be specified.


4.4  LABORATORY QA AND QC PROCEDURES

      The  QAPjP  should  describe  how  the following  required elements of the
laboratory QC program are to be implemented.

      4.4.1  Method  Proficiency

      Procedures should  be in place  for  demonstrating  proficiency  with each
analytical  method   routinely  used in the laboratory.    These should  include
procedures for demonstrating the precision and bias  of the method as performed
by  the  laboratory   and  procedures for  determining the method  detection  limit

                                   ONE - 18                         Revision 1
                                                                     July 1992

-------
(MDL).  All terminology, procedures and frequency of determinations associated
with the laboratory's establishment of the MDL and the reporting limit  should be
well-defined  and  well-documented.    Documented  precision,  bias,  and  MDL
information should be maintained  for  all methods performed in the laboratory.

      4.4.2  Control Limits

      Procedures should be in place for establishing and updating control limits
for  analysis.   Control  limits should  be  established  to  evaluate  laboratory
precision and bias based  on the analysis of control  samples.  Typically, control
limits for bias are  based  on the  historical  mean recovery plus or minus three
standard deviation units,  and  control  limits for precision range from zero (no
difference between duplicate control  samples)  to the historical mean relative
percent difference plus three standard deviation units.  Procedures should be in
place for monitoring  historical performance and should include graphical (control
charts) and/or tabular presentations  of the data.

      4.4.3  Laboratory Control Procedures

      Procedures should be in place for demonstrating  that the laboratory is in
control during each  data  collection  activity.   Analytical  data generated with
laboratory control samples that fall  within prescribed  limits are judged to be
generated while the  laboratory was in  control.  Data generated with laboratory
control samples that fall  outside the  established control limits are judged to
be generated during  an  "out-of-control"  situation.   These data are considered
suspect and should be repeated or reported with qualifiers.

      Laboratory  Control   Samples --    Laboratory control  samples should  be
      analyzed for each  analytical method  when  appropriate  for the method.   A
      laboratory control sample consists of either  a control matrix spiked with
      analytes representative  of  the  target  analytes  or a certified reference
      material.

      Laboratory control  sample(s) should be analyzed with each  batch of samples
      processed to verify that the precision  and  bias  of the analytical process
      are  within  control  limits.    The   results  of  the  laboratory  control
      sample(s) are  compared to control  limits  established for both precision
      and bias to determine  usability of the data.

      Method Blank -- When appropriate for the method,  a method  blank should be
      analyzed with  each  batch of samples processed  to assess contamination
      levels in the laboratory.  Guidelines should  be  in place for accepting or
      rejecting data based on  the level of contamination in the blank.

      Procedures should be in place for documenting the effect of the matrix on
method performance.  When  appropriate  for the method, there should be at least
one matrix spike and either  one matrix duplicate or one matrix spike duplicate
per analytical  batch. Additional  control samples may be necessary to assure data
quality to meet the  project-specific  DQOs.
                                   ONE - 19                         Revision 1
                                                                     July 1992

-------
      Matrix-Specific Bias --  Procedures  should  be  in place for determining the
      bias of the method  due  to the matrix.  These  procedures  should include
      preparation and analysis of matrix spikes,  selection and use of surrogates
      for organic methods, and  the method of  standard  additions for metal and
      inorganic methods.   When the concentration of the analyte in the sample is
      greater than 0.1%,  no spike is necessary.

      Matrix-Specific Precision  -- Procedures should be in place for determining
      the precision  of the method  for a specific matrix.   These  procedures
      should  include  analysis   of  matrix  duplicates  and/or  matrix  spike
      duplicates.  The frequency of use of these techniques should be based on
      the DQO for the data collection activity.

      Matrix-Specific  Detection Limit --  Procedures  should be in  place for
      determining the MDL  for a specific matrix type (e.g., wastewater treatment
      sludge, contaminated soil, etc).

      4.4.4  Deviations

      Any activity  not performed in accordance with  laboratory procedures or
QAPjPs is considered a deviation from plan.  All deviations from plan should be
documented as to the extent of,  and reason for,  the deviation.

      4.4.5  Corrective Action

      Errors, deficiencies, deviations, or laboratory events or data that fall
outside of  established acceptance criteria  should be investigated.   In some
instances, corrective  action may be  needed  to resolve  the problem and restore
proper functioning to the analytical system.  The  investigation of the problem
and any subsequent corrective action taken should  be documented.

      4.4.6  Data Handling

      Data resulting from  the analyses of samples should be reduced according to
protocols described  in the laboratory  procedures.   Computer programs used for
data reduction should be validated before use and  verified on a regular basis.
All information  used in  the  calculations (e.g., raw data,  calibration files,
tuning records, results of standard additions,  interference check results, and
blank- or background-correction  protocols) should be recorded in  order to enable
reconstruction  of the  final  result  at  a  later  date.    Information on the
preparation of the  sample  (e.g., weight  or  volume  of sample used,  percent dry
weight  for  solids,  extract  volume,  dilution  factor  used)   should  also  be
maintained in order  to  enable reconstruction of the  final result at a later date.

      All data should be reviewed by a second analyst or supervisor according to
laboratory  procedures  to  ensure that  calculations are  correct  and  to detect
transcription errors.  Spot checks  should be performed  on  computer calculations
to verify program validity.   Errors detected in the review process should be
referred to the  analyst(s) for  corrective action.   Data should  be reported in
accordance with  the  requirements of  the  end-user.   It is recommended that the
supporting documentation  include at a minimum:

                                   ONE - 20                         Revision 1
                                                                     July 1992

-------
      Laboratory name and address.

      Sample  information  (including  unique  sample  identification,  sample
      collection date and time,  date  of  sample  receipt,  and date(s) of sample
      preparation and analysis).

      Analytical results  reported with  an  appropriate number  of significant
      figures.

      Detection limits that  reflect dilutions, interferences, or correction for
      equivalent dry weight.

      Method reference.

      Appropriate QC results (correlation with sample  batch  should be traceable
      and documented).

      Data qualifiers with appropriate references and narrative on the quality
      of the results.
4.5  QUALITY ASSURANCE REVIEW

      The QA review consists of internal and external  assessments  to ensure that
QA/QC procedures are in use  and to ensure that laboratory staff conform to these
procedures.  QA review should be  conducted  as deemed appropriate  and necessary.


4.6  LABORATORY RECORDS

      Records provide the direct evidence and support for the necessary technical
interpretations,  judgements, and  discussions  concerning  project  activities.
These records, particularly those that are anticipated to be used  as evidentiary
data, should directly support technical  studies  and activities, and provide the
historical evidence  needed  for later  reviews  and analyses.  Records should be
legible,   identifiable,   and  retrievable,  and  protected   against   damage,
deterioration,  or  loss.     The  discussion  in  this   section (4.6)  outlines
recommended procedures for  record  keeping.   Organizations which  conduct field
sampling  should develop appropriate  record keeping procedures  which  satisfy
relevant technical and legal requirements.

      Laboratory records generally consist of bound notebooks with prenumbered
pages,  personnel  qualification  and training forms,  equipment maintenance and
calibration forms,  chain-of-custody forms,  sample analysis request  forms, and
analytical change request forms.  All records should be written in  indelible ink.

      Procedures for reviewing, approving, and revising laboratory  records should
be clearly  defined,  with  the lines of  authority  included.   Any  documentation
errors should be corrected  by drawing a  single line through the error so that it
remains legible and should be initialed by the responsible individual, along with
the date of change.  The correction is written adjacent to the error.

                                   ONE  - 21                         Revision 1
                                                                     July 1992

-------
      Strip-chart recorder printouts should be signed by the person who performed
the instrumental analysis.  If corrections need to be made in computerized data,
a system parallel to the corrections for handwritten data should be in place.

      Records of sample management should be available to permit the re-creation
of an analytical event for review in  the  case of  an audit  or investigation of a
dubious result.

      Laboratory records should include,  at least, the following:

      Operating Procedures -- Procedures should be available to those performing
      the  task outlined.   Any revisions  to laboratory  procedures  should  be
      written,  dated,  and distributed to  all affected  individuals  to ensure
      implementation  of changes.   Areas  covered by  operating  procedures are
      given in Sections 3.3  and 4.3.

      Quality Assurance Plans -- The QAPjP should be on file.

      Equipment Maintenance Documentation -- A history of the maintenance record
      of each  system serves as  an  indication of the  adequacy  of maintenance
      schedules and parts inventory.  As appropriate, the maintenance guidelines
      of the  equipment manufacturer  should be followed.   When  maintenance  is
      necessary,  it  should be  documented  in   either  standard  forms or  in
      logbooks.  Maintenance procedures  should  be clearly defined and written
      for each measurement  system and required support equipment.

      Proficiency -- Proficiency information  on all compounds  reported should be
      maintained and  should  include (1) precision; (2) bias; (3) method detec-
      tion limits;  (4) spike recovery, where applicable;  (5) surrogate recovery,
      where  applicable; (6) checks on reagent  purity, where  applicable; and
      (7) checks on glassware cleanliness, where applicable.

      Calibration Records & Traceability of Standards/Reagents --  Calibration is
      a  reproducible  reference point to which all sample measurements can  be
      correlated.   A sound  calibration  program  should  include  provisions for
      documenting frequency, conditions, standards, and records reflecting the
      calibration  history  of   a  measurement  system.   The  accuracy of the
      calibration standards  is  important because all  data will be in reference
      to  the  standards  used.    A  program for  verifying and  documenting the
      accuracy  and  traceability of all  working  standards against appropriate
      primary grade standards or the highest  quality standards available should
      be routinely  followed.

      Sample Management --All  required records  pertaining to sample management
      should  be maintained  and updated  regularly.   These  include chain-of-
      custody  forms,  sample  receipt forms, and sample disposition records.

      Original  Data  --  The raw data  and  calculated  results for  all  samples
      should be maintained in laboratory notebooks, logs,  benchsheets,  files or
      other  sample tracking  or data entry forms.   Instrumental output should be
      stored  in  a computer file or a hardcopy report.

                                   ONE -  22                         Revision  1
                                                                     July 1992

-------
      QC  Data  -- The  raw data and  calculated results  for all  QC  and field
      samples and standards should be maintained  in  the manner described  in the
      preceding  paragraph.   Documentation should allow  correlation  of  sample
      results with associated  QC  data.   Documentation should also include the
      source and lot  numbers  of standards for traceability.  QC samples include,
      but are not limited  to, control samples, method blanks, matrix spikes, and
      matrix spike duplicates.

      Correspondence -- Project correspondence can provide  evidence supporting
      technical interpretations.   Correspondence pertinent to the project  should
      be kept and placed  in the project files.

      Deviations -- All  deviations from procedural and planning documents  should
      be  recorded  in laboratory  notebooks.   Deviations  from QAPjPs  should be
      reviewed  and  approved  by   the  authorized   personnel  who  performed  the
      original technical  review or by their designees.

      Final Report -- A copy  of any  report issued  and  any  supporting documenta-
      tion should be retained.


5.0  DEFINITIONS

      The following  terms are defined for use in this document:

ACCURACY              The  closeness  of agreement  between an observed value and
                      an accepted reference  value.   When applied  to  a  set of
                      observed values,  accuracy  will  be  a combination of  a
                      random  component  and  of a  common  systematic  error  (or
                      bias) component.

BATCH:                A group of samples which behave  similarly with respect to
                      the  sampling or the testing procedures being  employed and
                      which are processed as a unit (see  Section  3.4.1 for field
                      samples and  Section 4.4.3 for laboratory samples).   For QC
                      purposes, if the number of  samples in a group is greater
                      than 20,  then each group of 20 samples or less will  all be
                      handled as  a separate  batch.

BIAS:                 The  deviation  due  to matrix effects  of the measured value
                      (x.  - xu) from a known spiked amount.  Bias  can be assessed
                      by comparing a  measured  value  to  an  accepted  reference
                      value in  a sample of known concentration or by determining
                      the  recovery of a  known amount of contaminant spiked into
                      a sample  (matrix spike).  Thus, the bias (B)  due to matrix
                      effects based  on a matrix spike  is  calculated as:

                                    B - (x.  -  xu  ) - K
                        where:
                                   ONE - 23                         Revision 1
                                                                     July 1992

-------
BLANK:

CONTROL SAMPLE:
DATA QUALITY
OBJECTIVES (DQOs):
DATA VALIDATION:
DUPLICATE:


EQUIPMENT BLANK:

EQUIPMENT RINSATE:
ESTIMATED
QUANTITATION
LIMIT (EQL):
                                    xt =  measured  value for spiked sample,
                                    xu = measured value for unspiked sample, and
                                    K  =  known value of the spike in the sample.

                      Using the following equation yields the percent recovery
              %R = 100  (x. - xu)/  K

see Equipment Rinsate,  Method Blank, Trip Blank.

A  QC  sample  introduced  into a  process  to  monitor the
performance of the system.

A  statement  of  the  overall level  of uncertainty that a
decision-maker  is willing to accept  in results derived
from environmental  data (see reference  2,  EPA/QAMS, July
16, 1986).   This  is  qualitatively distinct  from quality
measurements such as precision, bias, and detection limit.

The process of evaluating the available  data against the
project DQOs  to make  sure that the objectives are met.
Data  validation  may  be  very   rigorous,   or  cursory,
depending on  project DQOs.   The available data reviewed
will include analytical  results,  field QC data and lab QC
data, and may also include field records.
see  Matrix  Duplicate,
Duplicate.

see Equipment Rinsate.
Field  Duplicate,  Matrix  Spike
A  sample  of analyte-free media  which has  been used to
rinse  the sampling  equipment.    It  is  collected  after
completion of decontamination and prior to sampling.  This
blank is useful  in documenting adequate decontamination of
sampling equipment.

The  lowest  concentration that can  be reliably  achieved
within specified limits of precision  and  accuracy  during
routine  laboratory operating  conditions.   The  EQL is
generally  5  to 10  times the MDL.    However,  it may be
nominally chosen within these guidelines  to  simplify data
reporting.      For   many  analytes  the   EQL  analyte
concentration is selected as the  lowest non-zero standard
in the calibration curve.  Sample EQLs are highly matrix-
dependent.  The EQLs in  SW-846 are  provided  for guidance
and may not always be achievable.
                                   ONE  - 24
                                               Revision  1
                                                July  1992

-------
FIELD DUPLICATES:
LABORATORY CONTROL
SAMPLE:
MATRIX:


MATRIX DUPLICATE:


MATRIX SPIKE:
MATRIX SPIKE
DUPLICATES:
METHOD BLANK:
METHOD DETECTION
LIMIT (MDL):
Independent  samples  which  are  collected  as  close  as
possible to the  same  point  In space and time.  They  are
two separate samples taken from the same source,  stored in
separate containers,  and  analyzed independently.   These
duplicates are useful  in documenting the precision  of the^
sampling process.

A known matrix spiked with compound(s)  representative of
the target analytes.  This  is  used to document laboratory
performance.

The component or substrate  (e.g.,  surface water, drinking
water) which contains the analyte of interest.

An intralaboratory split sample which is used  to document
the precision of a method in a given sample matrix.

An aliquot of sample spiked  with a known concentration of
target analyte(s).   The spiking  occurs prior to  sample
preparation  and  analysis.    A  matrix  spike  is  used  to
document the bias of a method in a given sample matrix.

Intralaboratory  split  samples   spiked  with   identical
concentrations of target analyte(s).  The  spiking  occurs
prior to sample preparation and analysis.  They are  used
to document the precision and  bias of a  method in a given
sample matrix.

An analyte-free matrix to which all  reagents are added in
the  same  volumes  or  proportions  as   used  in   sample
processing.  The  method blank should be carried through
the complete sample preparation and analytical procedure.
The  method  blank  is  used  to  document  contamination
resulting from the analytical  process.

For  a  method  blank  to be  acceptable  for use  with   the
accompanying samples,  the concentration in the blank  of
any  analyte of  concern should not  be  higher  than   the
highest of either:

(l)The method detection limit, or

(2)Five percent of the regulatory  limit  for that analyte,
or
                      (3)Five
                      sample.
         percent  of  the  measured  concentration  in  the
The  minimum  concentration  of a  substance  that  can  be
measured and  reported with 99% confidence that  the analyte
concentration is greater than zero and  is determined from
                                   ONE - 25
                                              Revision 1
                                               July 1992

-------
analysis of a  sample  In  a given matrix type   containing
the analyte.

For  operational  purposes,  when  it   is   necessary  to
determine  the   MDL  in  the matrix,  the  MDL  should  be
determined by multiplying  the appropriate one-sided 99% t-
statistic  by   the  standard  deviation  obtained  from  a
minimum of three analyses  of a matrix spike containing  the
analyte of interest at a concentration three to five times
the estimated  MDL,  where the t-statistic is obtained from
standard references or the table below.
No. of samples;            t-statistic
      3                       6.96
      4                       4.54
      5                       3.75
      6                       3.36
      7                       3.14
      8                       3.00
      9                       2.90
     10                       2.82

Estimate the MDL as follows:
Obtain the concentration value that corresponds to:

a) an  instrument  signal/noise  ratio within the range  of
2.5 to 5.0, or

b) the region of  the standard  curve where  there is  a
significant change  in  sensitivity (i.e.,  a break  in the
slope of the standard curve).

Determine the variance (S2) for each analyte as follows:
                                  -"^2
where Xj * the  ith measurement of the  variable  x
and x = the average value of x;
              ONE -  26                         Revision  1
                                                July 1992

-------
ORGANIC-FREE
REAGENT WATER:
PRECISION:
                      Determine the standard deviation  (s) for each analyte  as
                      follows:
                                           2\1/2
               s = (S2)

Determine the MDL for each analyte as follows:
                                     MDL
                                           "
-------
PROJECT:
QUALITY ASSURANCE
PROJECT PLAN
(QAPJP):
RCRA:

REAGENT BLANK:

REAGENT GRADE:
REAGENT WATER:
REFERENCE MATERIAL:
SPLIT SAMPLES:
STANDARD ADDITION:
STANDARD CURVE:
                      where:
                       x = the arithmetic mean of the  Xj measurements, and S «
                      variance;  and the relative percent difference (RPD) when
                      only two samples are available.
               RPD = 100 [(x,  -
                                                              + x2)/2}].
Single or  multiple data collection  activities that are
related through the same planning sequence.

An orderly assemblage of detailed procedures designed to
produce  data  of  sufficient  quality  to  meet  the data
quality  objectives   for   a  specific  data   collection
activity.

The Resource Conservation and Recovery Act.

See Method Blank.

Analytical  reagent (AR) grade,  ACS  reagent  grade, and
reagent  grade  are  synonymous  terms  for  reagents  which
conform to the  current specifications of the Committee on
Analytical Reagents of the American Chemical Society.

Water that  has  been generated by any method which  would
achieve  the performance  specifications for ASTM Type II
water.    For  organic  analyses,   see  the  definition  of
organic-free reagent water.

A material containing known quantities of  target analytes
in solution or  in a homogeneous  matrix.   It  is used to
document the bias of the analytical process.

Aliquots  of sample  taken   from  the  same  container and
analyzed  independently.    In  cases  where  aliquots  of
samples are impossible to obtain, field duplicate samples
should be taken for the matrix duplicate analysis.   These
are usually taken after mixing or compositing and are used
to document intra- or interlaboratory precision.

The practice of adding  a known amount  of  an analyte to  a
sample  immediately prior to  analysis.   It is  typically
used to  evaluate  interferences.

A plot of concentrations of  known analyte standards versus
the  instrument  response  to  the  analyte.    Calibration
standards are prepared by successively diluting a standard
solution  to produce  working standards which  cover the
working  range  of  the instrument.   Standards  should be
prepared  at the  frequency  specified in the  appropriate
                                    ONE -  28
                                               Revision  1
                                                July 1992

-------
                      section.   The  calibration standards should  be prepared
                      using the same type  of acid or solvent  and  at the same
                      concentration as  will  result  in  the  samples  following
                      sample preparation.   This is applicable  to  organic and
                      inorganic chemical  analyses.

SURROGATE:            An  organic   compound  which  is  similar  to  the  target
                      analyte(s) in chemical  composition  and  behavior  in the
                      analytical process, but  which  is not  normally found in
                      environmental samples.

TRIP BLANK:           A sample  of analyte-free media  taken from the laboratory
                      to  the  sampling  site  and  returned  to  the  laboratory
                      unopened.   A  trip  blank is used to document contamination
                      attributable  to shipping  and field  handling  procedures.
                      This type of  blank is useful in documenting contamination
                      of volatile organics  samples.


6.0  REFERENCES

1.   Interim  Guidelines  and  Specifications  for  Preparing  Quality  Assurance
     Project Plans, QAMS-005/80, December 29, 1980, Office  of  Monitoring Systems
     and Quality Assurance, ORD, U.S.  EPA,  Washington,  DC  20460.

2.   Development of Data Quality Objectives,  Description of Stages I  and II, July
     16, 1986, Quality Assurance Management Staff,  ORD,  U.S. EPA, Washington, DC
     20460.

3.   RCRA  Ground-Water Monitoring Technical  Enforcement Guidance  Document,
     September, 1986,  Office of Waste  Programs Enforcement.   OSWER,  U.S. EPA,
     Washington, DC,  20460.

4.   DQO  Training  Software, Version  6.5,   December,  1988,  Quality  Assurance
     Management Staff, ORD, U.S. EPA,  Washington,  DC  20460.

5.   Preparing  Perfect Project  Plans,   EPA/600/9-89/087,   October  1989,  Risk
     Reduction Engineering Laboratory (Guy  Simes),  Cincinnati OH.

6.   ASTM Method D 1129-77, Specification for Reagent Water.   1991  Annual Book
     of ASTM Standards.  Volume 11.01  Water and Environmental Technology.

7.   Generation of Environmental  Data  Related  to Waste Management Activities
     (Draft).  February 1989.   ASTM.
                                   ONE - 29                         Revision 1
                                                                     July 1992

-------
                                     INDEX

Accuracy  1, 13, 22, 23", 24
Batch  12, 19, 21, 23"
Bias  2, 3, 17-20, 22, 23"-25, 28
Blank  11, 12, 14, 18-20, 23", 24, 25, 28, 29
   Equipment Rinsate  11, 12,  14, 18, 24*
   Method Blank  19, 24, 25", 28
   Reagent Blank  28"
   Trip Blank  12, 18, 24, 29*
Chain-of-Custody  9, 11, 13,  14,  18,  21,  22
Control Chart  18, 19
Control Sample  11, 12,  18,  19, 23, 24"
Data Quality Objectives  (DQO)  1-3, 8, 12, 19, 20, 24", 28
Decision-maker  2, 24
Duplicate  11, 12, 14, 18-20,  23, 24", 25, 27, 28
   Field Duplicate  11,  12,  24, 25", 28
   Matrix Duplicate  12, 19,  20,  24,  25", 28
   Matrix Spike Duplicate  12, 19, 20, 23, 24, 25"
Equipment Blank  11, 24*
Equipment Rinsate  11, 12, 14, 18, 24"
Estimated Quantitation Limit  (EQL)  24"
Field Duplicate  12, 24, 25", 28
Laboratory Control Sample  19, 25"
Matrix  11, 12, 18-20, 23-25", 26-28
Matrix Duplicate  12, 19, 20,  24, 25", 28
Matrix Spike  12, 18-20, 23,  25",  26,  27
Matrix Spike Duplicate   12,  19, 20, 23,  24, 25"
Method Blank  19, 24, 25",  28
Method Detection Limit (MDL)  18-20, 22,  24, 25*-27
Organic-Free Reagent Water   27",  28
Precision  1-3, 17-20, 22, 24, 25, 27",  28
Project  1-5, 7, 8, 11-14, 17-19, 21, 23, 24,  28"
Quality Assurance Project Plan (QAPjP)  1-9,  11,  12, 14, 15, 18, 20, 22, 23, 28"
RCRA  1, 8, 28"
Reagent Blank  28"
Reagent Grade  28"
Reagent Water  27, 28*
Reference Material  8, 11, 15, 18,  19, 28*
Split Samples  25, 28"
Standard Addition  20, 28"
Standard Curve  26, 28"
Surrogate  18, 20, 22, 29"
Trip Blank  12, 18, 24,  29"
   Definition of term.
                                   ONE  -  30                          Revision  1
                                                                      July 1992

-------
                            PREFACE AND OVERVIEW
PURPOSE OF THE MANUAL

     Test Methods for Evaluating Solid Waste (SW-846)  is intended to provide a
unified, up-to-date source of information  on sampling and analysis related to
compliance with RCRA regulations.   It  brings together into one reference all
sampling and testing methodology approved by the Office of Solid Waste for use
in implementing the RCRA regulatory  program.  The manual provides methodology
for collecting and testing representative samples of waste and other materials
to be monitored.  Aspects  of  sampling  and testing covered in SW-846 include
quality control, sampling  plan  development  and  implementation, analysis of
inorganic and  organic  constituents,  the  estimation  of  intrinsic physical
properties, and the appraisal of waste characteristics.

     The procedures described in this manual are meant to be comprehensive and
detailed, coupled  with  the  realization  that  the  problems  encountered in
sampling and analytical situations  require  a  certain amount of flexibility.
The solutions to these problems will  depend, in part, on the skill, training,
and experience of the analyst.    For  some  situations, it is possible to use
this manual  in  rote  fashion.    In  other  situations,  it  will  require a
combination of technical abilities, using  the  manual as guidance rather than
in a step-by-step, word-by-word fashion.    Although this puts an extra burden
on the  user,  it  is  unavoidable  because  of  the  variety  of sampling and
analytical conditions found with hazardous wastes.
ORGANIZATION AND FORMAT
     This manual  is divided into two  volumes.  Volume I focuses on laboratory
activities and  is divided  for  convenience  into  three  sections.  Volume IA
deals  with  quality  control,  selection  of  appropriate  test  methods, and
analytical methods for metallic species.    Volume  IB consists of methods for
organic  analytes.    Volume  1C  includes  a  variety  of  test  methods  for
miscellaneous   analytes  and  properties  for  use  in  evaluating  the  waste
characteristics.  Volume II deals with sample acquisition and includes quality
control, sampling plan design and  implementation, and field sampling methods.
Included for the  convenience  of  sampling  personnel  are discusssions of the
ground water, land treatment, and incineration monitoring regulations.

     Volume I begins with an overview  of the quality control precedures to be
imposed upon the  sampling and  analytical methods. The quality control chapter
(Chapter One) and the  methods  chapters  are  interdependent.  The analytical
procedures cannot be  used  without  a  thorough  understanding of the quality
control requirements and the means to  implement them.  This understanding can
be achieved only  be reviewing Chapter One and the analytical methods together.
It is expected  that  individual  laboratories,  using  SW-846 as the reference
                                  PREFACE -  1
                                                         Revision
                                                         Date  September 1986

-------
source, will  select  appropriate  methods  and  develop  a standard operating
procedure (SOP) to be followed by  the laboratory.   The SOP should incorporate
the pertinent information from this  manual   adopted to the specific needs and
circumstances of the individual laboratory as  well   as to the materials to be
evaluated.

     The method  selection  chapter  (Chapter  Two)   presents  a comprehensive
discussion of the application  of  these  methods  to  various matrices in the
determination of groups of analytes or specific analytes.  It aids the chemist
in constructing the correct  analytical  method  from  the array of procedures
which may  cover  the  matrix/analyte/concentration  combination of interests.
The  section  discusses  the  objective   of   the  testing  program  and  its
relationship to the choice of an analytical method.   Flow charts are presented
along with  tables  to  guide  in  the  selection  of  the  correct analytical
procedures to form the appropriate method.

     The analytical methods are  separated into distinct procedures describing
specific,   independent  analytical  operations.    These   include  extraction,
digestion, cleanup, and  determination.    This  format  allows linking of the
various steps  in the analysis  according   to:   the type of  sample  (e.g., water,
soil,  sludge,  still bottom);  analytes(s)  of  interest; needed  sensitivity; and
available analytical instrumentation.     The  chapters describing  Miscellaneous
Test Methods and   Properties,  however,   give complete  methods  which  are not
amenable  to  such segmentation  to form  discrete procedures.

     The  introductory material   at  the   beginning   of  each  section  containing
analytical   procedures    presents    information    on    sample   handling and
preservation,  safety, and  sample preparation.

     Part   II   of   Volume    I    (Chapters   Seven   and   Eight)  describes the
characteristics of a waste.    Sections   following  the  regulatory descriptions
contain  the methods  used  to   determine  if   the  waste  is  hazardous  because  it
exhibits  a  particular characteristic.

     Volume II gives background   information  on  statistical  and nonstatistical
aspects   of sampling.     It   also   presents practical   sampling  techniques
appropriate for situations presenting  a  variety  of  physical  conditions.

     A discussion   of   the  regulatory  requirements  with  respect to several
monitoring  categories  is  also given   in  this  volume.    These include ground
water  monitoring,  land   treatment,   and   incineration.     The  purpose of this
guidance  is to orient  the  user to  the  objective  of  the analysis,  and to assist
in developing  data quality objectives, sampling  plans,  and laboratory SOP's.

     Significant  interferences,  or  other  problems,  may  be encountered with
certain samples.   In these situations,  the  analyst 1s advised to contact the
Chief, Methods Section  (WH-562B)  Technical   Assessment Branch, Office of Solid
Waste, US EPA,  Washington,   DC     20460  (202-382-4761)  for assistance.  The
manual is intended to serve   all   those  with  a need to evaluate solid waste.
 Your comments, corrections,  suggestions, and questions concerning any material
 contained in,  or omitted  from,   this  manual  will be gratefully appreciated.
 Please direct  your comments  to the above address.


                                  PREFACE - 2
                                                          Revision      0
                                                          Date  September 1986

-------
                                 CHAPTER FOUR
                               ORGANIC ANALYTES
4.1   SAMPLING CONSIDERATIONS

      4.1.1  Introduction

      Following  the  initial and  critical  step  of designing  a  sampling plan
(Chapter Nine)  is  the  implementation of that plan  such  that a representative
sample of the solid waste is collected.  Once the sample has been collected it
must be stored  and preserved to maintain the chemical and physical properties
that it possessed at the time of collection.  The sample type,  type of containers
and their preparation,  possible  forms of contamination, and preservation methods
are  all  items  which  must  be   thoroughly  examined in  order to maintain the
integrity of the samples.   This  section  highlights  considerations which must be
addressed in order to maintain a  sample's integrity and representativeness.  This
section is, however,  applicable only to trace analyses.

      Quality  Control  (QC)  requirements need  not be  met  for  all  compounds
presented in the Table of Analytes for  the method in  use, rather, they must be
met  for all  compounds reported.    A   report  of  non-detect  is considered  a
quantitative report,  and must  meet  all  applicable QC  requirements  for that
compound and the method used.

      4.1.2  Sample Handling and Preservation

      This  section deals  separately  with volatile  and semivolatile organics.
Refer to Chapter Two and Table 4-1 of this section for  sample containers, sample
preservation, and sample holding time information.

      Volatile Organics

      Standard 40 ml  glass screw-cap  VOA vials with  Teflon lined silicone septa
may be used for both liquid and solid matrices.  The  vials and septa should be
washed with  soap and water and rinsed  with  distilled deionized  water.   After
thoroughly cleaning the vials and  septa,  they  should  be  placed in  an oven and
dried at 100°C  for  approximately one  hour.

NOTE: Do not heat the  septa for extended periods of time (i.e.,  more than one
      hour, because the silicone begins to slowly degrade at 105°C).

      When collecting the samples,  liquids and  solids  should be introduced into
the vials gently to reduce agitation which might drive off volatile compounds.
In general,  liquid samples should be poured into the vial without introducing any
air bubbles within the vial as  it is being filled.  Should bubbling occur as a
result of violent pouring,  the  sample must be poured out  and the vial refilled.
The vials should be completely  filled at the  time of sampling, so that when the
septum cap is fitted and sealed, and the vial  inverted, no headspace is visible.
The sample should be hermetically sealed in the vial  at the time  of sampling, and
must not be opened prior to analysis  to preserve their integrity.
                                   FOUR - 1                         Revision 2
                                                                September 1994

-------
            due to  differing  solubility and diffusion properties  of gases in
            LIQUID matrices at different  temperatures,  it  is possible for the
            sample to generate some  headspace  during  storage.   This headspace
            will appear in the form of micro bubbles,  and should not invalidate
            a sample for volatiles analysis.

            The presence of a  macro bubble in a sample vial generally indicates
            either  improper  sampling technique or  a  source of  gas evolution
            within the  sample.   The  latter case  is  usually accompanied  by a
            buildup  of  pressure  within  the vial,  (e.g.  carbonate-containing
            samples  preserved with  acid).    Studies  conducted  by  the  USEPA
            (EMSL-Ci, unpublished data) indicate that "pea-sized" bubbles (i.e.,
            bubbles  not exceeding  1/4  inch  or  6 mm  in  diameter) did  not
            adversely  affect  volatiles  data.   These bubbles were generally
            encountered in wastewater samples,  which  are more  susceptible to
            variations in gas solubility than  are groundwater samples.

 At the time of analysis,  the aliquot to be analyzed  should be  taken from the
vial with a gas-tight syringe  inserted directly through the  septum of the vial.
Only one analytical  sample  can be taken from each  vial.  If these guidelines are
not followed,  the validity of  the data generated from  the samples is suspect.

      VGA vials for  samples with solid or semi-solid  matrices  (e.g., sludges)
should be completely  filled as best as possible.   The  vials should be  tapped
slightly as they  are filled  to  try and  eliminate  as much  free air  space as
possible.  Two vials should also be filled per sample  location.

      At least two  VGA  vials  should be filled and  labeled  immediately  at the
point at which the sample is collected. They should NOT be  filled near a running
motor or any  type of exhaust system because  discharged fumes and  vapors may
contaminate the samples.  The  two vials from each  sampling location  should then
be  sealed  in  separate plastic  bags  to  prevent  cross-contamination  between
samples, particularly if the sampled waste is suspected  of containing  high levels
of volatile organics.   (Activated  carbon may  also be included in  the bags to
prevent cross-contamination from  highly contaminated samples).  VOA  samples may
also be contaminated by  diffusion of volatile organics  through the septum during
shipment and storage.  To monitor possible contamination, a trip blank prepared
from organic-free reagent  water  (as  defined in Chapter  One) should be carried
throughout the sampling, storage, and shipping process.

      Semivolatile Organics (including Pesticides, PCBs and Herbicides.)

      Containers used to collect samples for the determination of semivolatile
organic  compounds  should   be  soap  and water washed  followed by methanol  (or
isopropanol) rinsing  (see  Sec.  4.1.4 for specific instructions on glassware
cleaning).   The sample containers should  be  of glass or Teflon, and  have screw-
caps with  Teflon  lined septa.   In situations  where  Teflon  is  not available,
solvent-rinsed aluminum foil   may be used as a  liner.  However,  acidic or basic
samples may react with the  aluminum foil,  causing  eventual contamination of the
sample.  Plastic containers or lids may NOT be used for the  storage of samples
due to  the  possibility of sample  contamination from  the  phthalate esters and
other hydrocarbons within the  plastic.  Sample containers should be  filled with
care so as  to prevent any portion of the collected  sample coming in contact with

                                   FOUR - 2                         Revision 2
                                                                September 1994

-------
the  sampler's gloves,  thus  causing  contamination.   Samples  should  not  be
collected or  stored in the  presence of  exhaust  fumes.   If the sample comes in
contact with the sampler (e.g. if an automatic sampler is used), run organic-free
reagent water through the sampler and use as a field blank.

      4.1.3   Safety

      Safety  should always  be the primary consideration  in  the  collection of
samples.  A thorough understanding of the waste production process, as well as
all of the potential hazards making up the waste,  should be  investigated whenever
possible.   The site  should be visually evaluated  just  prior to  sampling to
determine additional safety measures.   Minimum protection of gloves and safety
glasses  should be  worn  to  prevent sample  contact  with the skin and  eyes.   A
respirator  should  be  worn  even  when  working outdoors if organic  vapors are
present.  More hazardous sampling missions may require the use of supplied air
and special  clothing.

      4.1.4   Cleaning of Glassware

      In the analysis  of samples  containing components  in  the parts per billion
range, the preparation of  scrupulously clean glassware is  necessary.  Failure to
do  so  can  lead to  a  myriad  of  problems in  the  interpretation  of  the final
chromatograms  due  to  the  presence   of   extraneous  peaks  resulting  from
contamination.  Particular  care  must  be taken with  glassware  such  as Soxhlet
extractors,  Kuderna-Danish evaporative concentrators, sampling-train components,
or any other glassware coming  in contact with an  extract that will be evaporated
to a smaller volume.  The  process  of concentrating the  compounds of interest in
this operation may similarly concentrate the contaminating substance(s), which
may seriously distort the results.

      The basic cleaning steps are:

      1.  Removal  of surface residuals immediately after use;

      2.  Hot soak to loosen and float most particulate material;

      3.  Hot water rinse to flush away floated particulates;

      4.  Soak with an oxidizing agent to destroy traces of organic compounds;

      5.  Hot water rinse  to flush away materials loosened  by the deep penetrant
          soak;

      6.  Distilled water rinse to remove metallic deposits from the tap water;

      7.  Alcohol,  e.g.,  isopropanol  or methanol,  rinse to flush  off any final
          traces  of organic materials  and remove the water; and

      8.  Flushing the item  immediately before use with some of the same solvent
          that will  be used in the analysis.
                                   FOUR - 3                         Revision 2
                                                                September 1994

-------
      Each of these eight fundamental steps are discussed here in the order in
which they appeared on the preceeding page.

      1.  As soon possible after  glassware  (i.e.,  beakers,  pipets,  flasks, or
          bottles) has come in contact with sample or standards, the glassware
          should  be   flushed  with  alcohol  before  it  is  placed  in the  hot
          detergent soak.   If this  is  not done,  the  soak bath may serve to
          contaminate all other glassware placed therein.

      2.  The hot soak consists of  a  bath  of  a suitable detergent  in water of
          50°C  or higher.  The detergent, powder or liquid, should  be entirely
          synthetic and not a fatty  acid base.   There are very few areas of the
          country where  the  water hardness is sufficiently low to  avoid  the
          formation of some hard-water scum  resulting from the reaction between
          calcium and magnesium salts with a fatty acid soap.  This hard-water
          scum or curd would  have  an  affinity particularly for many chlorinated
          compounds and,  being almost wholly water-insoluble, would deposit on
          all glassware  in the bath in a thin  film.

          There are many  suitable  detergents on the wholesale and retail market.
          Most of the  common  liquid  dishwashing detergents sold at retail  are
          satisfactory but  are more expensive  than other  comparable products
          sold industrially.   Alconox, in powder or tablet form, is manufactured
          by Alconox,  Inc., New York,  and is marketed by  a number of laboratory
          supply firms.  Sparkleen,  another  powdered product, is distributed by
          Fisher Scientific Company.

      3.  No comments required.

      4.  The most common  and highly  effective oxidizing agent  for removal of
          traces of organic compounds is the traditional chromic acid solution
          made  up of concentrated  sulfuric  acid  and potassium   or  sodium
          dichromate.   For maximum efficiency, the soak solution should be hot
          (40-50°C).    Safety precautions  must  be  rigidly  observed  in  the
          handling of this solution.   Prescribed safety  gear  should include
          safety goggles, rubber gloves, and apron.  The bench area where this
          operation is conducted  should  be  covered with fluorocarbon sheeting
          because spattering  will  disintegrate any unprotected surfaces.

          The potential hazards of using chromic-sulfuric acid mixture are great
          and have been  well  publicized.  There are now commercially available
          substitutes that possess the advantage of safety in handling.  These
          are biodegradable concentrates with  a claimed  cleaning strength equal
          to the chromic acid solution.   They  are alkaline, equivalent to ca.
          0.1  N  NaOH   upon  dilution,  and are  claimed  to  remove dried blood,
          silicone greases, distillation residues, insoluble organic residues,
          etc.  They are further  claimed to remove radioactive traces and will
          not attack glass or exert  a corrosive effect on skin or clothing.  One
          such product is "Chem Solv 2157," manufactured  by Mallinckrodt and
          available through laboratory supply firms. Another comparable product
          is "Detex," a  product of Borer-Chemie, Solothurn, Switzerland.
                                   FOUR - 4                         Revision 2
                                                                September 1994

-------
      5, 6, and 7.  No comments required.

      8.  There is always  a possibility that between the time of washing and the
          next use, the glassware could pick up some contamination from either
          the  air or  direct  contact.   To ensure  against this,  it  is  good
          practice to  flush the  item immediately before use with  some of the
          same solvent that will  be used in the analysis.

      The drying and storage of the cleaned glassware is of critical importance
to  prevent  the   beneficial effects   of  the  scrupulous cleaning  from  being
nullified.    Pegboard  drying   is  not  recommended.    It  is  recommended  that
laboratory glassware and  equipment be dried at  100°C.   Under  no circumstances
should such small  items be left  in the open without  protective  covering.   The
dust  cloud raised  by  the daily  sweeping of  the laboratory  floor  can  most
effectively recontaminate the  clean glassware.

      As  an  alternate  to solvent  rinsing,  the glassware  can  be heated  to  a
minimum of 300°C  to  vaporize  any organics.   Do not  use  this  high temperature
treatment  on  volumetric  glassware,  glassware  with  ground glass joints,  or
sintered glassware.

      4.1.5  High Concentration Samples

            Cross contamination of trace concentration  samples  may occur  when
      prepared in the same laboratory with high  concentration samples.  Ideally,
      if  both type  samples   are  being  handled,  a  laboratory  and  glassware
      dedicated solely  to  the preparation of high concentration samples would be
      available for this  purpose.  If this  is  not  feasible, as  a  minimum  when
      preparing high concentration  samples, disposable glassware should be  used
      or,  at  least,  glassware dedicated entirely to  the high  concentration
      samples.    Avoid  cleaning   glassware  used  for  both  trace  and  high
      concentration samples in the same area.
                                   FOUR -  5                         Revision  2
                                                                September 1994

-------













t/}
1 1 )
v_
~~
1
'
ca
*™
o
o
1C
0

^
rO
>
i_
0)
00
0)
D-







&-
a>
c

















00
rO
^~
0
a>
+•*
^•j
^
c
)
re
•o

2
















0
0
^"

Vt
r"
o
o

OO
00 C
(0 O
0)
x: h-
O 4->
E —
a> 2
-o
•£ 1- -0
C r—

^E ^0 T3
in c c
CM O ••-
^H U i—


t/1
a>
r^
o.
E
re
oo

Q)
re
oo 2^
O
•— -o
fO ^j
o 15
i- ^.
CD 4->
a, §
•— u
•JT C
^ O
re >
re
•o

«s-
1— 1
















•^
o
^>

*.
^^
o
0
o


^r
•r»
(/I
fO
>
	 1
E

O
•«f
X

CM






O)
c

c
o
o
00
a> •—
•— re
1 *
TO «^
00 00
a)
•o cc
•^
3 0
CT JC
•^
_l





































00
a.
re
u
E
3
*J
Q.
(U
00

c
*r*

c
o
fZI
O)
K-
















C
0)
00
O)
i-
CL,









00
>>
re
"O

«*•
1— 1


1 ^^
QJ O
S- «
—I CL
E l*-
C 0
LO QJ
CM O) 00
rH XI O.
O
re oo i.
re -a
C £Z.
JZ
a> <_> -C
Q.X: ••-
E 2 2
ra
cu ci)
*J C >
0 ••- S-

r~ 4^ OO
r— C QJ
0 0 i-
c_> u a.
00
a.
re
u
+•> E
••- 3
X +J
a.
oo ai
r— 00
re
••- T3
C
	 1 •(-
E r—

0 C
* 0
X ^
Q)
CM 1—

4->
C

00
0)

Q_

O)
C
0

x:

"^
3
T3
•r"
00
O)
a:
















•o-
c •
• re r—
C re
O CO "-
•r- 1— >
-t-> a.
3 E <:
r— re O
o oo •>
00
X —1
O) -r- E
4-> E
rO O
<*- O *
Lo n- **
O I- O
-C 5 O

a> •*
E >,t'-
3 f*~ OO ^
•i- +J C P—
TJ C re O
o a> i- o
oo O +•> O












































00
>.,
re
^^

*




c ^
•
^^

"
0
O
u

in
^.
:r
a.
o


j *
(/)
3
-o
3
00
Q.
re
U
f E
•i- 3
2 •*->
a.
00 O)
i — 00
re
••- TJ
c
	 1 •!-
E r—

O C
^ o
x "4!
a>
CM h-











a>
•u i—

00 L>
c ••-
0) O
^H. ^^
2£
O 0
<: <:









(/)
>>
re
^3

i-H
















^
o
^^

•>.
r«_
o
o
o

00
00 .C
re -t->
CT 3:
-C ^O
^ Ci)
3 i—
O re
T3
2 0)
c
_i -^ E
E re 3
LO C Q.
CM O fl)
i— i O 00


00
a>
C^
•o
3
^->
OO

•o
re
00
+^
c
O!
*f»
•D
QJ
00
r—
•r~
O
oo






































•
C3*
OO
31
re
z
•o
•^
'o
oo
s_
o

J"T
Z

«
OO
fM
z:

.c
+->
•r-
2

CM
V
Z
a.

4->
t/)
3
•1-5
•a


r*






CM
C
O
•^
oo
•r"
>
CD
OL


























UD
O
| i
U^





































SJ
2

L.
O)

E
OJ

a.
O)
































































-------

























-^^
cu
J
^
.r—
-4-*
C
O
CJ

**
I
J"s*

UJ
-1
CQ
^t
1 	



















































cu
•f-
i_
^
D>
C
**"*
"O
'o
1C










cu

>r_
[ ^
re
^.
cu
co
CU

Q-
















S-
OJ
c
.,-_
(TJ

c:
o
0



















CO
CO
(TJ
r—
O

cu
4-3
>}
r—
itJ
C"
^^




cu
4-3
co
3


to
CU
Q.
E
(TJ
CO













CO
CU
-o

(J

_Q
S-
0)


•a
E


to
CO
O
Q_

co
cu

• r—
CJ

I ^
CO
cu
Q-

OJ
E

i_
o
H—
-C
o
o
c
(TS
a
s_
o
*^-*
CO
O
• i—
C
(TJ
a
s_
o

cu
1 —
•^
4-3
(TJ
r-~
O
>
•r—
E
O)
co


cu
c
o





CO
to
(TJ
CD


r- 1 O

C 4J
•r- U C
.C (TJ -I- C7>
4-> i- -E c
•r- 4-3 4-> -i—
s x-.- s •
cu 3 o c
-a r— o
cu ~o "o i — •!—
+J c CU O 4-3
0 (TJ N 4- O
fTJ >> fO
S- to i — to S-
-M >, (TS >,+J
X (TJ C (TJ X
CU T3  -r-
1 ^ >r_
co S
E "O

CO 4->
CU U
a. s_
E -t->
(TJ X
CO O)











o
o
^
•X
r—
O
0
o


s-
o
•
r-^
(TJ CO
O) CO
1 (TJ
in r-

0
to «d-
4->
O C
(TJ •(— en
4_> 4J -r-
x -r- s •
Q) 3 O C
i— O
^3 ^D r~ *r—
C CD O 4-J
 i — tO J-
>, RS >,4->
(TJ C (TJ X
T3 (TJ T3 O>
























C
0
•4!
cu
_C CU • Cnl—
4-J
3
O

cu

>r-
^

1
E

IT)
CXJ




co
cu

a.

(TJ


cu

CO
(TJ
3

-^
cu

(TS
s_
I *
E
Q)
o

o
o




C
•r—
P_

C
o

t^_
cu
^~

.c
4->
• r—
3
















co
CU
^~
cx

(TJ
co

s-
CD
+j
fTJ
3




C3
s_
X 
CO C
CU 0)
a; to
0)
o s-
Z Q-







x:
4_>
•r-
3t

S- T3
CU -r-
C i—
•r—
(TJ -O
4-> CU
C C
O •!-
U r—


































,^

C
O) •>-
xi -c
^j *r-»
to S
3
e -o
cu
CO 4->
CU U
i— (TJ
a. s-
E-M
(TJ X
CO 0)
cu "
-t-> ^1-
'='0
CO o
-C
4->
3
"O ^~"
•a o
 r—

0
CO •*
4->
O E
(TJ T- cn
S- -C C
_!__* ^_) •(—
x ••- s •
cu 2 o c
r— O
T3 -0 r— •—
C CU O 4->
(TJ tsl 4- CJ
to i — to i-
(TJ C « X
T3 rtJ -O O)
























C
O
<+I
cu

•— i O
co ^t-
cu -i- u c
JD -C ns T- cn
4J S_ J= C
+J *i~~ 4-J -|-J *r~
to S X -i- S •
3 CU S 0 C
E -0 r- 0
CU T3 "O r— ••-
CO 4-> C CU O 4->
CU O (TJ N 4- O
r^ (TJ >> (TS
CX S- to i — tO t-
E 4J >, (TS >>4->
(TJ X (TS C (TS X
CO CU -O (TJ "O CU











C_3
o
"*
«*
r~—
O
o
o




CO
to c
(TJ O
"0)4!
cu
• cni— ^= 1—
o
i-
X O)
JQ
CNJ E
(TJ
s_
o -
	 1
• 1
1 	 f-H
(TJ
01 X
1
1— ( ^J-


1 ^
c
cu
to
cu
s-
O-

cu
c

^_
o

c-
o

p^
(TJ
3
•^
• ^
to
cu
a:







r~
4-3
•r—
3

S- "O
cu -f-
C i —
•r-
nj -a
4-1 CU
C £=
O -r-
CJ i —

































4-3
^ <"
O 4->
E ••-
CU S
"O
•i- S- -0
3 cu •.-
E ^~
_J -r-
E « -0
4_) CU
O C C
in o -i-
CSJ O i—



co
cu
cn

3
t—
CO

"^J
c
(TJ

CO
4_3
E
CU
E
•i—
-a
cu
CO
^^
r—
• ^"
0
CO

























,
cu
to
3

-o
'ai

t^

0
4-3
c
o
si
cx

^^
o

(0

o
-O
(TJ


CU
.c:


E


"O
CD
E
S-
o
4-
S-
CD
Q.

CL)
(~i

>^
(TJ
E

E
0

4-*
03
>
s-
cu
to
cu
s_
Q.
i
CU
i_
Q-

c\
C\J «d"
C 
O •—!
co S-
•T- CU
> -Q
CU E
r^ Q;
4J
CX
CU
CO





















i^^^

i
y^
^
O
1 1























































-------
4.2  SAMPLE PREPARATION METHODS

     4.2.1  EXTRACTIONS AND PREPARATIONS

         The following methods are included in this section:

         Method 3500A:     Organic Extraction and Sample Preparation
         Method 3510B:     Separatory Funnel Liquid-Liquid Extraction
         Method 3520B:     Continuous Liquid-Liquid Extraction
         Method 3540B:     Soxhlet Extraction
         Method 3541:      Automated Soxhlet Extraction
         Method 3550A:     Ultrasonic Extraction
         Method 3580A:     Waste Dilution
         Method 5030A:     Purge-and-Trap
         Method 5040A:     Analysis  of  Sorbent  Cartridges  from  Volatile
                           Organic    Sampling    Train    (VOST):        Gas
                           Chromatography/Mass Spectrometry Technique
         Method 5041:      Protocol for Analysis of Sorbent Cartridges from
                           Volatile Organic  Sampling  Train  (VOST):   Wide-
                           bore Capillary Column Technique
                                  FOUR -  8                       Revision 2
                                                             September 1994

-------
                                 METHOD 3500A

                   ORGANIC  EXTRACTION AND  SAMPLE  PREPARATION
1.0   SCOPE AND APPLICATION

      1.1   The  3500 Methods  are  procedures  for  quantitatively  extracting
nonvolatile and  semivolatile  organic compounds from various  sample matrices.
Cleanup and/or analysis  of the  resultant  extracts are described in Chapter Two,
Sections 2.3.2 and 2.3.1, respectively.

      1.2   Method 3580  describes a solvent dilution technique that may be used
on non-aqueous  nonvolatile and semivolatile organic samples  prior  to cleanup
and/or analysis.

      1.3   The  5000  Methods  are procedures for preparing  samples  containing
volatile organic compounds for quantitative analysis.

      1.4   Refer to the specific method of interest for further details.


2.0    SUMMARY OF METHOD

      2.1   3500 Methods:   A  sample  of a known volume or weight  is solvent
extracted.  The resultant extract is dried and then concentrated in a Kuderna-
Danish apparatus (if necessary).   Other concentration devices or techniques may
be used  in place  of the  Kuderna-Danish  concentrator  if the quality control
requirements of the determinative methods are met (Method 8000,  Section 8.0).

      2.2   5000 Methods:  Refer to the specific method of interest.


3.0   INTERFERENCES

      3.1   Samples  requiring  analysis for volatile organic  compounds,  can be
contaminated by diffusion of volatile organics  (particularly chlorofluoro-carbons
and methylene chloride)  through the sample container septum during shipment and
storage.   A  field  blank prepared from organic-free reagent  water  and carried
through sampling and subsequent storage and handling can serve as  a check on such
contamination.

      3.2   Solvents, reagents, glassware,  and  other sample processing hardware
may yield artifacts  and/or interferences to sample analysis. All these materials
must be demonstrated to be free from interferences  under the conditions of the
analysis  by  analyzing  method  blanks.    Specific  selection  of reagents  and
purification of solvents by distillation in all-glass  systems may be required.
Refer to Chapter One for specific guidance on quality  control procedures.

      3.3   Interferences  coextracted  from the samples  will  vary considerably
from source to source.  If analysis of an extracted sample is prevented due to
interferences, further cleanup  of the sample extract may be necessary.  Refer to
Method 3600 for guidance on cleanup procedures.


                                   3500A  -  1                         Revision 1
                                                                     July 1992

-------
      3.4   Phthalate esters contaminate many types of products commonly found
in the laboratory.   Plastics,  in particular, must be avoided because phthalates
are  commonly used  as  plasticizers  and  are  easily  extracted  from  plastic
materials.  Serious  phthalate contamination may result at  any time if consistent
quality control  is not practiced.

      3.5   Glassware contamination  resulting in  analyte  degradation:   Soap
residue on glassware may cause degradation of certain analytes.  Specifically,
aldrin, heptachlor, and most organophosphorus  pesticides will  degrade  in this
situation.   This problem  is especially pronounced with  glassware  that may be
difficult to rinse (e.g.,  500 ml K-D  flask).  These items should be hand-rinsed
very carefully to avoid this problem.


4.0   APPARATUS AND MATERIALS

      4.1   Refer to the  specific method  of  interest  for a description of the
apparatus and materials needed.


5.0   REAGENTS

      5.1   Refer to the  specific method  of  interest  for a description of the
solvents needed.

      5.2   Organic-free reagent water.  All  references to water in this method
refer to organic-free reagent water as defined in Chapter One.

      5.3   Stock standards: Stock solutions may be prepared from pure standard
materials or purchased as certified solutions.

            5.3.1 Purgeable  stock  standards:    Prepare  stock  standards  in
      methanol using assayed liquids or gases, as appropriate.  Because of the
      toxicity  of  some  of the  organohalides,   primary  dilutions  of  these
      materials should be prepared in a hood.

                  5.3.1.1     Place  about 9.8  ml  of  methanol  in  a 10-mL tared
            ground-glass-stoppered volumetric flask.   Allow the flask to stand,
            unstoppered,  for about 10 min or until all alcohol-wetted surfaces
            have dried.   Weigh the flask to the nearest  0.0001 g.

                  5.3.1.2     Using  a  100-/itL  syringe, immediately  add two or
            more drops of assayed reference material to the  flask, then reweigh.
            The liquid  must fall directly into the alcohol without contacting
            the neck of the flask.

                  5.3.1.3     Reweigh, dilute  to  volume, stopper,  then mix by
             inverting the flask several times.  Calculate the  concentration in
            milligrams  per  liter (mg/L)  from the net gain  in weight.   When
            compound purity is assayed to be 96% or greater, the weight may be
             used without correction to calculate  the  concentration  of the stock
             standard.  Commercially prepared stock standards may be  used at any
            concentration if  they  are certified by the  manufacturer  or by an
             independent source.

                                   3500A - 2                        Revision  1
                                                                     July  1992

-------
                  5.3.1.4     Transfer  the  stock  standard  solution  into  a
            Teflon-sealed screw-cap bottle.  Store, with minimal  headspace, at -
            10°C to -20°C  and  protect  from  light.

                  5.3.1.5     All standards must be replaced after 1 month,  or
            sooner if comparison with check standards indicates a problem.

            5.3.2 Semivolatile  stock  standards:   Base/neutral  and  acid  stock
      standards are prepared  in methanol.   Organochlorine pesticide standards
      are prepared in acetone.

                  5.3.2.1     Stock  standard solutions  should  be   stored  in
            Teflon-sealed containers  at  4°C.  The  solutions  should  be checked
            frequently for stability.  These  solutions  must  be replaced after
            six months,  or sooner  if comparison  with  quality  control  check
            samples indicate a  problem.

      5.4   Surrogate standards: A surrogate standard (i.e.,  a chemically inert
compound not expected to  occur  in  an  environmental  sample)  should be added to
each  sample,  blank,  and matrix spike  sample  just  prior  to  extraction  or
processing.   The  recovery of the  surrogate  standard  is  used  to monitor for
unusual matrix effects, gross  sample processing errors,  etc.   Surrogate recovery
is evaluated for  acceptance by  determining whether the measured concentration
falls within the acceptance limits.  Recommended surrogates  for different analyte
groups follow.   However, these compounds,  or others that better correspond to the
analyte group, may be used for other analyte  groups  as well.  Normally three or
more standards are added for each analyte group.

            5.4.1 Base/neutral  and  acid  surrogate  spiking  solutions:    The
      following are recommended surrogate standards.

            Base/neutral                   Acid

            2-Fluorobiphenyl              2-Fluorophenol
            Ni trobenzene-d5                2,4,6-Tri bromophenol
            Terphenyl-du                  Phenol-d6

                  5.4.1.1     Prepare a surrogate standard spiking solution in
            methanol that contains the base/neutral compounds  at  a concentration
            of  100  mg/L, and  the  acid  compounds  at 200  mg/L  for  water and
            sediment/soil samples (low- and medium-level).   For waste samples,
            the concentration  should be 500 mg/L for base/neutrals and 1000 mg/L
            for acids.

            5.4.2 Organochlorine pesticide/PCB surrogate spiking solution:  The
      following   are   recommended   surrogate   standards   for  Organochlorine
      pesticides/PCBs.

            Orqanochlorine pesticides/PCBs

            Dibutylchlorendate  (DBC)  (if available)
            2,4,5,6-Tetrachloro-meta-xylene (TCMX)
                                   3500A -  3                         Revision 1
                                                                     July 1992

-------
                  5.4.2.1     Prepare a surrogate standard spiking solution at
            a concentration  of  1  mg/L in acetone for  water  and sediment/soil
            samples.  For waste samples, the concentration should be 5 mg/L.

            5.4.3 Purgeable  surrogate spiking  solution:    The  following  are
      recommended surrogate standards for volatile organics.

            Purqeable organics

            p-Bromof1uorobenzene
            l,2-Dichloroethane-d4
            Toluene-d8

                  5.4.3.1     Prepare a surrogate spiking solution (as described
            in  Section  5.3.1   or  through  secondary  dilution  of  the  stock
            standard)  in methanol  containing  the  surrogate  standards at  a
            concentration of 25 mg/L.

      5.5   Matrix  spike standards:   Select  five  or more analytes  from each
analyte group  for use in a  spiking  solution.  The  following  are recommended
matrix spike standard mixtures  for a few analyte  groups.   These compounds, or
others that better correspond to the analyte  group, may be used for other analyte
groups as well.

            5.5.1 Base/neutral  and  acid matrix spiking solution:   Prepare a
      spiking  solution  in  methanol  that   contains   each   of  the  following
      base/neutral compounds at 100 mg/L and the acid compounds at 200 mg/L for
      water and  sediment/soil  samples.   The  concentration of  these compounds
      should be five times higher for waste samples.

            Base/neutrals                 Acids

            1,2,4-Tri chlorobenzene        Pentachlorophenol
            Acenaphthene                  Phenol
            2,4-Dinitrotoluene            2-Chlorophenol
            Pyrene                        4-Chloro-3-methylphenol
            N-Nitroso-di-n-propylamine    4-Nitrophenol
            1,4-Dichlorobenzene

            5.5.2 Organochlorine pesticide matrix spiking solution:  Prepare a
      spiking  solution  in  acetone  or  methanol that  contains  the  following
      pesticides  in  the  concentrations  specified  for water and  sediment/soil.
      The concentration  should  be five times higher for waste  samples.

            Pesticide                     Concentration (mq/L)

            Lindane                             0.2
            Heptachlor                          0.2
            Aldrin                              0.2
            Dieldrin                            0.5
            Endrin                              0.5
            4,4'-DDT                            0.5
                                   3500A - 4                        Revision  1
                                                                     July  1992

-------
            5.5.3 Purgeable matrix spiking solution:  Prepare a spiking solution
      in methanol  that  contains the following compounds at  a  concentration of 25
      mg/L.

            Purgeable organics

            1,1-Dichloroethene
            Trichloroethene
            Chlorobenzene
            Toluene
            Benzene
6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory  material  to  the  Organic  Analyte  Chapter,
Section 4.1.
7.0   PROCEDURE

      7.1   Semivolatile  organic sample  extraction:    Water,  soil/sediment,
sludge,  and  waste  samples  requiring  analysis  for  base/neutral  and  acid
extractables and/or  organochlorine pesticides must  undergo  solvent extraction
prior to analysis.   This manual  contains four methods that may be used for this
purpose:  Method 3510;  Method 3520; Method  3540;  and Method 3550.   The method
that should  be used on a particular sample, is highly dependent upon the physical
characteristics of that sample.   Therefore,  review these four methods prior to
choosing one in particular.   Appropriate surrogate standards  and, if necessary,
matrix spiking solutions are added to the sample prior to extraction  for all four
methods.

            7.1.1 Method 3510:  Applicable to the extraction and concentration
      of  water-insoluble and  slightly  water-soluble   organics  from  aqueous
      samples.   A  measured volume of  sample  is  solvent  extracted  using  a
      separatory funnel.  The extract  is dried, concentrated  and, if necessary,
      exchanged into a  solvent compatible with  further analysis.   Method 3520
      should be used if an  emulsion  forms  between  the solvent-sample  phases,
      which can not be broken up by mechanical  techniques.

            7.1.2 Method 3520:  Applicable to the extraction and concentration
      of  water-insoluble and  slightly  water-soluble   organics  from  aqueous
      samples.  A measured volume of sample is extracted with  an organic solvent
      in a continuous liquid-liquid extractor.  The solvent  must have a density
      greater than that of the sample.  The  extract  is dried, concentrated and,
      if necessary, exchanged into a  solvent compatible with further analysis.
      The limitations of Method 3510 concerning solvent-sample phase separation
      do not interfere with  this procedure.

            7.1.3 Method 3540:  This  is  a procedure for extracting nonvolatile
      and semivolatile organic compounds from solids such as  soils, sludges, and
      wastes.   A  solid  sample is mixed  with anhydrous  sodium  sulfate, placed
      into  an  extraction thimble or between  two  plugs  of glass  wool,  and
      extracted using  an appropriate  solvent   in  a Soxhlet extractor.   The

                                  3500A - 5                        Revision 1
                                                                     July 1992

-------
      extract is dried, concentrated and, if necessary, exchanged into a solvent
      compatible with further analysis.

            7.1.4 Method 3550:  This method is applicable to the  extraction of
      nonvolatile and semivolatile organic  compounds from solids such as soils,
      sludges, and wastes  using  the technique of ultrasonic extraction.   Two
      procedures  are  detailed depending  upon the  expected concentration  of
      organics  in  the sample; a  low concentration and  a  high  concentration
      method.  In both,  a known weight of sample is mixed with anhydrous sodium
      sulfate and solvent extracted using ultrasonic extraction.  The extract is
      dried, concentrated and, if necessary, exchanged  into a solvent compatible
      with further analysis.

            7.1.5 Method 3580:  This method describes  the technique of solvent
      dilution of non-aqueous waste samples.  It is  designed  for wastes that may
      contain organic chemicals at a level  greater  than  20,000 mg/kg  and that
      are soluble in  the dilution  solvent.   When using this method, the analyst
      must use  caution  in  determining  the  correct concentration  of  spike and
      surrogate solution to avoid diluting out these compounds when diluting the
      sample.  The loss of  surrogate and spike data  should only occur in samples
      containing a high concentration of analytes which  is unknown at the time
      of  extraction  or  where sample  interferences could  not  be  eliminated
      following the best attempts at extract cleanup .by  the laboratory.

      7.2   Volatile organic  sample preparation:  There  are three methods for
volatile sample preparation:  Method 5030;  Method  5040;  and direct injection.
Method 5030  is  the most widely applicable  procedure for  analysis of volatile
organics, while the direct  injection technique may have limited applicability to
aqueous matrices.

            7.2.1 Method 5030:  This method describes the  technique  of purge-
      and-trap  for  the   introduction  of  purgeable  organics   into  a  gas
      chromatograph.   This procedure is  applicable for use with aqueous samples
      directly  and  to   solids,  wastes,  soils/sediments,   and  water-miscible
      liquids  following  appropriate preparation.   An   inert  gas is  bubbled
      through the sample, which will efficiently transfer  the purgeable organics
      from  the  aqueous  phase to  the  vapor phase.  The  vapor phase  is  swept
      through a sorbent  trap where the purgeables are trapped.  After purging is
      completed, the  trap is heated and backflushed with the  inert  gas to desorb
      the purgeables  onto a gas chromatographic column.  Prior to application of
      the purge-and-trap procedure, all  samples (including blanks, spikes, and
      duplicates)  should be  spiked  with surrogate  standards and,  if required,
      with matrix  spiking  compounds.

            7.2.2  Method 5040:  This method is applicable to the  investigation
      of sorbent cartridges  from volatile organic sampling train  (VOST).

      7.3   Sample analysis:  Following preparation of a sample  by one of the
methods described  above, the sample is ready for  further  analysis.  For samples
requiring volatile organic analysis, application of one of the methods described
above is followed directly by gas  chromatographic analysis (Methods 8010,  8011,
8015,  8020,  8021, 8030,  8240 and 8260).   Samples prepared  for semivolatile
analysis  may,  if necessary,  undergo  cleanup  (See  Method  3600)  prior  to
application of a specific  determinative method.

                                   3500A -  6                        Revision  1
                                                                     July 1992

-------
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One for  specific guidance  on quality  control
procedures.

      8.2   Before processing any samples,  the analyst should demonstrate through
the  analysis  of a  reagent water  blank  that all  glassware  and  reagents  are
interference free.  Each time a set of samples is processed,  a method blank(s)
should be processed as a safeguard  against chronic laboratory contamination.  The
blank samples should be  carried through all stages of the sample preparation and
measurement.

      8.3   Surrogate standards should be added to all samples when specified in
the appropriate determinative method in Chapter Four, Section 4.3

      8.4   A reagent blank,  a matrix  spike,  and a  duplicate or matrix spike
duplicate must  be  performed  for each analytical batch (up to a  maximum of 20
samples) analyzed.

      8.5   For GC or GC/MS analysis, the  analytical  system performance must be
verified by analyzing quality control  (QC) check samples.  Method 8000,  Section
8.0 discusses in detail  the process of verification;  however,  preparation of the
QC check sample concentrate is dependent upon the method being evaluated.

            8.5.1 Volatile  organic  QC   check   samples:     QC   check   sample
      concentrates containing each analyte of interest are spiked into  reagent
      water  (defined  as the QC check  sample)  and  analyzed  by  purge-and-trap
      (Method 5030).  The concentration of each analyte in the QC check sample
      is 20 ng/L.  The evaluation of system performance is discussed in detail
      in Method 8000, beginning with Paragraph 8.6

            8.5.2 Semi volatile  organic  QC  check samples:    To  evaluate  the
      performance of the analytical method, the QC check samples must be handled
      in exactly the same manner as actual  samples.  Therefore, 1.0 ml of the QC
      check  sample  concentrate is spiked  into each  of  four 1-L aliquots of
      reagent water (now  called  the QC  check  sample),  extracted, and  then
      analyzed  by  GC.   The  variety of  semivolatile analytes  which   may be
      analyzed  by  GC is such  that the concentration  of  the QC  check sample
      concentrate is different  for the different analytical techniques presented
      in the manual.  Method 8000 discusses in detail  the procedure of verifying
      the detection  system once the QC check  sample has  been prepared.   The
      concentrations of the QC  check  sample concentrate for the various  methods
      are as follows:

                  8.5.2.1     Method 8040 -  Phenols:   The  QC  check  sample
            concentrate  should contain each analyte  at  a  concentration of 100
            mg/L in 2-propanol.

                  8.5.2.2     Method 8060  -  Phthalate esters:   The QC check
            sample  concentrate should  contain  the  following analytes  at  the
            following concentrations in acetone:  butyl  benzyl  phthalate, 10
            mg/L; bis(2-ethylhexyl) phthalate,  50 mg/L; di-n-octylphthalate, 50
            mg/L; and any other phthalate  at 25 mg/L.

                                   3500A - 7                         Revision 1
                                                                     July 1992

-------
      8.5.2.3     Method 8070 - Nitrosamines:  The QC check sample
concentrate should contain each analyte  at  20  mg/L in methanol  or
some other water miscible solvent.

      8.5.2.4     Method 8080 - Orqanochlorine pesticides and PCBs;
The QC check sample concentrate should contain each single-component
analyte at  the  following concentrations in acetone  or  some other
water  miscible  solvent:   4,4'-DDD,  10  mg/L;  4,4'-DDT,  10 mg/L;
endosulfan  II,  10 mg/L; endosulfan  sulfate,  10 mg/L;  endrin,  10
mg/L; and any other single-component pesticide at 2 mg/L .  If the
method is only to be  used to  analyze  PCBs, chlordane, or toxaphene,
the   QC   check  sample  concentrate  should  contain   the  most
representative multicomponent  parameter  at a concentration of  50
mg/L in acetone.

      8.5.2.5     Method 8090 - Nitroaromatics and Cyclic Ketones:
The QC check sample concentrate should contain each analyte at the
following concentrations in  acetone:   each dinitrotoluene  at  20
mg/L; and isophorone and nitrobenzene at 100 mg/L.

      8.5.2.6     Method 8100 - Polynuclear aromatic hydrocarbons;
The QC check sample concentrate should contain each analyte at the
following concentrations in  acetonitrile:   naphthalene,  100 mg/L;
acenaphthylene,  100  mg/L;  acenaphthene,  100 mg/L;  fluorene,  100
mg/L;    phenanthrene,    100   mg/L;   anthracene,    100    mg/L;
benzo(k)fluoranthene, 5 mg/L; and any other PAH at 10 mg/L  .

      8.3.2.7     Method 8110  -  Haloethers;  The  QC check sample
concentrate  should contain  each analyte   at  a  concentration  of
20 mg/L in methanol or some other water miscible solvent.

      8.5.2.8     Method 8120  - Chlorinated hydrocarbons:   The QC
check  sample  concentrate   should  contain   each   analyte   at  the
following   concentrations    in    acetone:   hexachloro-substituted
hydrocarbons, 10 mg/L;  and  any other chlorinated hydrocarbon, 100
mg/L.

      8.3.2.9     Method 8140/8141   -  Organophosphorus  compounds:
The  QC check  sample  concentrate should  contain  each  analyte  in
acetone at  a  concentration  1,000  times more concentrated than the
selected spike concentration.

      8.3.2.10    Method 8150  -   Chlorinated  herbicides:   The  QC
check  sample concentrate should contain each analyte in acetone at
a  concentration 1,000  times  more concentrated than  the selected
spike concentration.

      8.3.2.11    Method 8250/8270 - Semi volatile organics:  The QC
check  sample concentrate should contain each analyte in acetone at
a concentration of 100 mg/L.

      8.3.2.12    Method 8310 - Polynuclear aromatic hydrocarbons:
The QC check sample concentrate should contain each analyte at the
following concentrations in  acetonitrile:   naphthalene, 100 mg/L;

                       3500A -  8                         Revision  1
                                                         July 1992

-------
            acenaphthylene,  100  mg/L;  acenaphthene,   100   mg/L;   fluorene,
            100 mg/L;   phenanthrene,    100   mg/L;   anthracene,   100   mg/L;
            benzo(k)fluoranthene, 5 mg/L; and any other PAH at 10 mg/L.


9.0   METHOD PERFORMANCE

      9.1   The recovery of surrogate standards is used to monitor unusual matrix
effects,  sample processing  problems,  etc.    The recovery  of  matrix  spiking
compounds indicates the presence or absence of unusual  matrix effects.

      9.2   The performance  of this method  will be  dictated by the  overall
performance  of the  sample  preparation in  combination  with the  analytical
determinative method.
10.0  REFERENCES

      10.1  None required.
                                  3500A  - 9                         Revision 1
                                                                     July 1992

-------
                                         METHOD 3500A
                        ORGANIC EXTRACTION AND  SAMPLE PREPARATION
7  1 1 Method
   3510
                            Non- aqueous
Semi volatile
Organics
                            No
                      713 Method
                         3510
                        C1eanup,
                      Determina 11v
                        Method
Yes
) 	 >

7 1 5 Method
3580



7 1 4 Method
3550

\,



                   So rbent
                   Ca r tridges
                   from VOST
                                                                                      722 Method
                                                                                         5030
                                    Direct
                                    Injection
                                           3500A  -  10
                               Revision  1
                                July  1992

-------
                                 METHOD 3510B

                  SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION


1.0   SCOPE AND APPLICATION

      1.1   This method describes  a  procedure  for isolating organic compounds
           is  samples.   The  method also  describes  concentration  techniques
           r  preparing  the extract for the  appropriate determinative methods
           n Co*-  A 7 f\-f f""h !in4"a v» Cnnv*
      _,_   ..--_ 	— __  _____ ____  _.  r
from  aqueous  samples.   The  method al<
suitable for  preparing  the extract for
described in Sec. 4.3 of Chapter Four.
      1.2   This method  is applicable to  the isolation  and  concentration of
water-insoluble and slightly water-soluble organics in preparation for a variety
of chromatographic procedures.


2.0   SUMMARY OF METHOD

      2.1   A measured volume  of sample, usually 1 liter, at a  specified pH  (see
Table 1),  is serially  extracted with methylene  chloride using  a  separatory
funnel.   The extract is dried, concentrated (if necessary), and, as  necessary,
exchanged into a solvent compatible with the cleanup or  determinative method to
be used (see Table 1 for appropriate exchange solvents).


3.0   INTERFERENCES

      3.1   Refer to Method 3500.

      3.2   Under basic extraction conditions  required to  separate  analytes for
the packed columns of Method 8250, the decomposition of some analytes has been
demonstrated.  Organochlorine  pesticides may dechlorinate,  phthalate  esters may
exchange, and phenols may react to form tannates.   These  reactions increase with
increasing pH,  and  are decreased  by  the  shorter reaction  times  available in
Method 3510.   Methods 3520/8270, 3510/8270,  and  3510/8250,  respectively,  are
preferred over Method 3520/8250 for the analysis of these  classes of  compounds.


4.0   APPARATUS AND MATERIALS

      4.1   Separatory funnel  - 2 liter,  with Teflon stopcock.

      4.2   Drying column  - 20 mm ID Pyrex chromatographic column with Pyrex
glass wool at bottom and a Teflon stopcock.

      NOTE: Fritted  glass  discs are  difficult  to decontaminate  after  highly
            contaminated extracts  have been  passed through.   Columns without
            frits may  be  purchased.   Use  a  small  pad of  Pyrex  glass wool to
            retain the  adsorbent.   Prewash the glass wool pad with 50  mL of
            acetone followed by  50 mL of elution  solvent  prior to packing the
            column with adsorbent.
                                   3510B  -  1                         Revision 2
                                                                September 1994

-------
      4.3   Kuderna-Danish (K-D) apparatus.

            4.3.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
      equivalent).  A  ground-glass stopper is used to  prevent  evaporation of
      extracts.

            4.3.2 Evaporation    flask  -    500  ml  (Kontes  K-570001-500  or
      equivalent).   Attach  to  concentrator  tube  with  springs,  clamps,  or
      equivalent.

            4.3.3 Snyder  column  -  Three  ball  macro (Kontes  K-503000-0121  or
      equivalent).

            4.3.4 Snyder  column -  Two ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.3.5 Springs -  1/2 inch (Kontes  K-662750 or equivalent).

      4.4   Boiling chips - Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.5   Water  bath  - Heated,  with  concentric  ring  cover,  capable  of
temperature control (±5°C).   The bath  should be  used in  a hood.

      4.6   Vials - 2 ml, glass with Teflon lined screw-caps or crimp tops.

      4.7   pH indicator paper - pH range including the desired extraction pH.

      4.8   Erlenmeyer flask - 250 ml.

      4.9   Syringe - 5 ml.

      4.10  Graduated cylinder - 1 liter.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall  be used in all  tests. Unless otherwise
indicated, it is intended that all  reagents shall conform  to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such specifications are available. Other grades may be used, provided it is first
ascertained that the reagent  is  of sufficiently  high  purity to permit its use
without lessening the accuracy of the determination. Reagents should be stored
in glass to prevent the leaching of contaminants from plastic containers.

      5.2   Organic-free reagent water - All references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Sodium  hydroxide solution  (ION),  NaOH.   Dissolve 40 g NaOH  in
organic-free reagent water and dilute to  100 ml.

      5.4   Sodium sulfate (granular,  anhydrous), Na2S04.  Purify by heating at
400°C for 4 hours in a shallow tray, or by precleaning the sodium sulfate with


                                   3510B  - 2                         Revision 2
                                                                September 1994

-------
methylene chloride.  If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed,  demonstrating that there is no  interference from
the sodium sulfate.

      5.5   Sulfuric acid solution (1:1 v/v),  H2S04.   Slowly add  50 ml  of H2S04
(sp. gr. 1.84) to 50 ml of organic-free reagent water.

      5.6   Extraction/exchange solvents

            5.6.1 Methylene chloride, CH2C12 - Pesticide quality or equivalent.

            5.6.2 Hexane, C6H14  -  Pesticide quality or  equivalent.

            5.6.3 2-Propanol, CH3CH(OH)CH3  - Pesticide quality or equivalent.

            5.6.4 Cyclohexane, C6H12  -  Pesticide quality or equivalent.

            5.6.5 Acetonitrile,  CH3CN - Pesticide  quality or  equivalent.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Sec. 4.1.


7.0   PROCEDURE

      7.1   Using a  1  liter  graduated cylinder, measure  1 liter (nominal) of
sample  and transfer  it  quantitatively to  the  separatory   funnel.   If high
concentrations are anticipated, a  smaller  volume  may  be  used  and then diluted
with  organic-free  reagent water  to  1 liter.   Add  1.0  mL  of  the surrogate
standards  to   all  samples,   spikes,   and  blanks   (see Method  3500   and  the
determinative method to be used,  for  details on the surrogate  standard  solution
and the matrix spike solution).  For the sample in each  analytical  batch  selected
for spiking, add 1.0 mL of the matrix  spiking standard.  For base/neutral-acid
analysis, the  amount  added of the surrogates and  matrix spiking compounds  should
result in a final concentration of 100 ng/juL  of each  base/neutral analyte and
200 ng//uL of  each acid  analyte  in  the  extract to  be analyzed  (assuming a 1 /nL
injection). If Method 3640, Gel-Permeation  Cleanup, is  to be used, add twice the
volume of surrogates  and matrix spiking compounds since half the extract is lost
due to loading of the GPC column.

      7.2   Check the   pH  of the sample  with  wide-range  pH  paper  and,  if
necessary,  adjust  the  pH  to  that  indicated  in Table  1   for   the   specific
determinative  method that will be used to analyze  the  extract.

      7.3   Add 60 mL of methylene chloride to the separatory funnel.

      7.4   Seal  and  shake the separatory funnel  vigorously for 1-2 minutes with
periodic venting to  release excess pressure.
                                   3510B  -  3                         Revision 2
                                                                September 1994

-------
      NOTE: Methylene  chloride   creates   excessive   pressure  very  rapidly;
            therefore, initial  venting should  be  done  immediately  after the
            separatory funnel has been sealed and shaken once.  Venting of the
            separatory funnel should be into a hood to avoid needless exposure
            of the analyst to solvent vapors.

      7.5   Allow  the  organic layer  to  separate from  the  water phase  for  a
minimum of 10 minutes.   If  the  emulsion  interface  between layers is more than
one-third the  size of the  solvent  layer,  the analyst  must  employ  mechanical
techniques to complete the phase separation.  The optimum technique depends upon
the sample and may  include  stirring,  filtration  of  the emulsion through glass
wool, centrifugation, or  other physical methods.  Collect the  solvent extract in
an Erlenmeyer flask.   If  the  emulsion cannot be broken (recovery of < 80% of the
methylene chloride, corrected for the water solubility of methylene chloride),
transfer the  sample,  solvent, and  emulsion  into the extraction chamber  of  a
continuous extractor and  proceed as described in Method  3520,  Continuous Liquid-
Liquid Extraction.

      7.6   Repeat the extraction  two more times using fresh portions of solvent
(Sees. 7.3 through 7.5).   Combine the three solvent extracts.

      7.7   If further pH adjustment and extraction is required, adjust the pH
of the aqueous phase to the desired pH indicated in Table 1.   Serially extract
three  times  with  60  mL of methylene  chloride,  as  outlined  in  Sees.  7.3
through 7.5.   Collect  and combine  the  extracts  and  label the combined extract
appropriately.

      7.8   If performing GC/MS analysis (Method 8270), the acid/neutral and base
extracts may be combined prior to concentration.  However, in some situations,
separate concentration and analysis  of the acid/neutral  and base  extracts may be
preferable (e.g.  if for regulatory purposes the presence  or absence of specific
acid/neutral  or base compounds at low concentrations must be determined, separate
extract analyses may be warranted).

      7.9   Perform the  concentration  (if  necessary)  using  the Kuderna-Danish
(K-D) Technique (Sees. 7.10.1 through 7.10.4).

      7.10  K-D Technique

            7.10.1      Assemble   a  Kuderna-Danish    (K-D)   concentrator  by
      attaching a  10 mL  concentrator  tube  to  a  500 mL  evaporation  flask.   Dry
      the extract by passing  it through a drying column containing about 10 cm
      of  anhydrous  sodium   sulfate.    Collect   the  dried  extract   in  a  K-D
      concentrator.   Rinse  the  Erlenmeyer flask, which  contained  the solvent
      extract, with 20-30 mL of methylene chloride and add it to the column to
      complete the quantitative transfer.

            7.10.2      Add  one or two clean  boiling  chips  to  the  flask and
      attach a three  ball  Snyder  column.   Prewet the  Snyder column by adding
      about 1 mL of methylene chloride to  the  top of  the  column.  Place the K-D
      apparatus on  a hot water bath  (15-20°C  above  the boiling point  of the
      solvent) so that the  concentrator tube  is  partially immersed  in the hot
      water and the entire lower rounded surface of the flask  is  bathed with hot


                                   3510B  -  4                         Revision  2
                                                                September 1994

-------
      vapor.   Adjust the  vertical  position  of the  apparatus  and  the  water
      temperature as required to complete  the  concentration  in  10-20 minutes.
      At the proper rate of distillation the balls of the column will actively
      chatter, but the  chambers  will  not flood.  When  the apparent  volume of
      liquid reaches  1  ml,  remove  the K-D  apparatus  from the water bath and
      allow it to drain and cool  for at least 10 minutes.

            7.10.3      If a solvent exchange is required (as  indicated in Table
      1), momentarily  remove the  Snyder column,  add 50 ml of  the exchange
      solvent, a new boiling chip, and  reattach  the Snyder column.  Concentrate
      the extract, as described  in  Sec.  7.11,  raising  the temperature  of the
      water bath, if necessary,  to  maintain proper distillation.

            7.10.4      Remove the Snyder  column  and  rinse  the  flask  and its
      lower joints into  the  concentrator  tube with 1-2 ml of methylene chloride
      or  exchange solvent.    If sulfur  crystals  are  a  problem, proceed  to
      Method 3660 for cleanup.  The extract may be further concentrated by using
      the technique  outlined in Sec.  7.11 or  adjusted to  10.0 ml with the
      solvent last used.

      7.11  If further concentration  is indicated in Table 1,  either the micro-
Snyder column technique  (7.11.1) or nitrogen blowdown technique (7.11.2) is used
to adjust the extract to the final  volume required.

            7.11.1      Micro-Snyder Column Technique

                  7.11.1.1    If further  concentration is indicated in Table 1,
            add another clean boiling chip to the concentrator tube and attach
            a two ball micro-Snyder column.  Prewet the column by adding 0.5 ml
            of methylene chloride or  exchange  solvent  to the  top of the column.
            Place the K-D  apparatus in a hot water bath so that the concentrator
            tube is partially immersed in the hot water.  Adjust the vertical
            position of  the  apparatus and the water temperature, as required, to
            complete the concentration in 5-10 minutes.  At  the  proper rate of
            distillation the balls of the column will actively chatter,  but the
            chambers will  not flood.  When the  apparent volume of liquid reaches
            0.5 ml,  remove the K-D apparatus from the water bath and allow it to
            drain and cool  for at least  10 minutes.   Remove  the  Snyder  column
            and rinse the  flask and its lower joints into the  concentrator tube
            with 0.2 ml  of extraction  solvent.   Adjust the final  volume to 1.0-
            2.0 ml,  as indicated  in  Table 1,  with  solvent.

            7.11.2      Nitrogen  Blowdown Technique

                  7.11.2.1    Place  the concentrator tube in  a warm bath (35°C)
            and evaporate  the solvent  volume to  0.5 ml using a gentle stream of
            clean, dry nitrogen (filtered through a column of activated carbon).

                  CAUTION:    New plastic tubing must not be  used  between the
                              carbon   trap  and  the   sample,  since  it   may
                              introduce interferences.
                                  3510B - 5                         Revision  2
                                                                September 1994

-------
                  7.11.2.2    The internal wall  of the tube must be rinsed down
            several times with methylene  chloride  or appropriate solvent during
            the operation.  During evaporation,  the tube solvent level must be
            positioned to avoid water  condensation.   Under  normal  procedures,
            the extract must not be allowed to become dry.

                  CAUTION:    When the volume of solvent  is reduced below 1 ml,
                              semivolatile analytes may be lost.

      7.12  The extract may now be  analyzed  for the  target  analytes using the
appropriate determinative  technique(s)  (see  Sec.  4.3 of  this Chapter).   If
analysis  of  the   extract  will  not  be   performed immediately,  stopper  the
concentrator tube and store refrigerated.  If the  extract will  be stored longer
than 2 days it should be  transferred to a vial with a Teflon lined screw-cap or
crimp top, and labeled appropriately.


8.0   QUALITY CONTROL

      8.1   Any reagent blanks or matrix spike  samples  should be subjected to
exactly the same analytical procedures as those used on actual samples.

      8.2   Refer  to  Chapter  One  for  specific  quality  control  procedures and
Method 3500 for extraction and sample preparation procedures.


9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods for performance data.


10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing  Test  Procedures for the
      Analysis of  Pollutants Under the Clean Water Act; Final Rule and Interim
      Final Rule and Proposed Rule," October 26,  1984.
                                   3510B - 6                        Revision 2
                                                                September  1994

-------
   oo
   o
   O
   o

   CO
   ^
   o
   •—t
   QL
   rx
LLJ O
_i u.
co
«=C oo
   O
   o
   t—
   X
  Q.
  OO



4->
re S-
c -4-*
•r- X

(J
rO
i-
Q) +->
E X
3 CU
'o 4-
> O






QJ
cn-t->
c c
rO QJ
JC >
O r—
X O
LjJ 00






CU
cn -i-3
c c
rO CU
-C >
CJ r—
x o
LU 00
































CU
3
'o o


^y
CU
s_
•r—
3
QJ O
S- <4-






^5
CU
S-
• r—
3
CT S-
QJ O
S- M-






^3
cu
s-
• r—
3
CT S-
QJ 0
S- <4-
C
>> 0
S- -r-
ro +->
•o o
C ro
0 S-
O +•>
QJ X
OO QJ

C
o
r— -4-i
ro U
•r- ro
+-> S-
•i- +->
C X
i— i QJ
CU
rO
C
i
S-
QJ
.+J
CU
o
'~-
£
•— '
oo
S
rO
C.
	 1
^


CL
3
C
ro
r^
O







CL
3
C
ro
QJ
r—
0






00
•r—
00
>>

"ro
C
ro








=C
CL








~T"
a.



-a
o
.c
4_)
CU
E
n
O
o
1 — t
o o
—i O






o o
i— I CV1









cu cu
c c
rO ro
X X
CU QJ
•^ •*=





^
C
ro
CL CU
O C
S- ro
a. x
i cu







cu cu
c c
o o
c c
-o
cu
>
cu
(J
QJ
S-

CXI 00
V\ rO





O O
«d- IO
O O
CO 00




0 0
O O






O O
CXI CXI

0)
-o
S-
o
.c
o
cu
c
cu
QJ r—
C >•
rO .C
X +J
QJ CU
-^ E






(—
0
cu c
C rO
ro s:
X +->
QJ QJ







CU CU
C C
O O
C C
T3 TJ
CU CU
> >
CU CU
u o
cu cu
s- s-

00 00
rO rO





— i 0
10 r-.
O 0
00 00




O 0
0 0






0 0
O 0
i— 1 r— 1








CU OJ
c c
ro ro
X X
cu cu
-*= -^








cu cu
c c
rO ro
X X
QJ CU







QJ CU
C C
o o
c c







CTI cn
i i
in LO





O i— i
00 00
o o
00 00




O 0
i— 1 i-H






o o
CNJ CM





cu
ro
X
cu
QJ .C
c o
rO r—
X 0
QJ >i
.C U








CU
C
re CU
X C
cu o







QJ CU
c c
O 0
c c
^
QJ
>
cu
o
cu
s-
O"*
I 00
in ro





0 0
ai o
O i— <
00 00




0 O
O r-H






0 O
C\J CSI









cu cu
c c
ro re
X X
cu cu
-*= -c








QJ CU
c c
rO ro
X X
CU QJ







cu cu
c c
0 O
c c
^
cu cu
> >
CU QJ
0 U
cu cu
S- S-

00 00
ro ro





0 O
•— 1 CSJ
r-l i— 1
00 00




0 0
r-l O






O 0
CXI O
i-H








QJ QJ
c c
rO ro
X X
cu cu
•= -=








cu cu
c c
re ro
X X
cu cu







QJ CU
c c
0 O
c c
-o
cu
>
cu
o
QJ
s-
00
00 I
re ID





— i O
cxj «a-
I— 1 1— I
OO 00




0 0
O — i






0
o
1 t 1








a>
c
ro
X
m
-*= '








a>
c
ro CU
X C
QJ O
-C C






QJ
c
O CXJ
C V
T5
QJ
>
cu
o
QJ
S-
l» 4
00 i— 1
ro A




u
.a
"i 0
** in
•— < CXI
OO 00




o o
1— t 1— 1








1 1













1 1



QJ
1—
•^~
S-
+->
• r—
c
o
cu +->
c cu
o u
C rO






cu
i— i C
<-* 0
A C
-o
cu
>
*r~
cu
o
cu
s-

CXI 00
V ro




D
a
0 0
r- •— t
CXI CO
OO OO

• — -
S-
-o
0 0
-H O






o
o
1 1— 1
eu
-a
S-
o
0
cu
c
Q)
r—
>^
-C
4-)
QJ
1 E
cu
si
o
JC.
u
QJ
r— C
0 QJ
C r-
ro >,
J=. -C.
4-> +->
cu cu
E E






QJ QJ
c c
o o
c c
•O TJ
CU OJ
> >
•r™ *i~
cu cu
u o
CU QJ
i. S-

00 00
rO ro





r-l O
CXI t— l
CO «3-
00 00
                                                                                ro
                                                                                C
                                                                                O
                                                                                a.
                                                                                o
                                                                                ro
                                                                                ro
      Q.
      3
      C
      rO
      QJ

       JQ
                           OJ E
                          OL O)
                             4->
                             Q.
                             QJ
                                                                                s
      re
      T3
 SQ 3
 " o 01
                                                                               o <-> o    o
                                                                               *f O 4-
                                                                               °-      §'
                                                                                             QJ
 o "d oo


 QJ >> o


   C "QJ
  • ro ^
Q

C5«
 O -t-J
 ro X  •
 S- QJ >i
-I-J    S-
 X QJ rO
 QJ C 00
   ro oo
r— v^ QJ
 O QJ O
 C .JE QJ
 rO    C
 Q	1 C
 0E3

 Q-O oo
 i  I-H -t-J
CXI    (J
^ ro ro

 E C +->

o- S
                                                                                          si
                                                                                          is
                                                                                             ro
              ro
           S- S-
           3 -M
           U 3
           o ai
           o c
 O rd

 oo -o

'o'u
 C ro
 CU
   QJ
                                                                                          (J
                                                                                             ro
                                                                                             «**>
                          CO
                          O
                                                                                                         in
                                                                                                         ro
                                                                                     0)
ii-
T3 -c
 QJ Q-
S j_
 >>,?
                                                                                     Q)
                                                                               0 S
                                                                               f^ O
        a; £  s- r-

      *""* 13 e^_Q fO
      g O-Q  ^^
     /^ s-  -  ^ e
     t/O       = *-»
     __    (/)  CZ %—>


     o^S's
     ^-s£°
     **- J3 ^  3 00
      ° ro QJ  ^ oo
       t— +j  QJ O
      >>-i- ro  oo r—
     +-> ro -—
     •r: > ro ni QJ
      
                                                                               r— ro
                                                                               O >
                                                                               C -r-
                                                                               QJ i-
        00 jc  C

      0^^°-^
      Q> 3 M_ -t->
      Q--O O  O +->
      oo QJ    ra i—
        O oo  S- 3
      QJ O oo +J oo
        S- O  X QJ
             -U S.
                                                                               O_ T3 I— Q	1
                                                                               ro
                                                                                          o -o

-------
                           METHOD 3510B
         SEPARATORY  FUNNEL LIQUID-LIQUID EXTRACTION
j   Start   J
     IT
7.1 Add surrogate
 standards to all
 samples,  spikes,
   and blanks.
                                7.7 Collect
                               and combine
                             extracts and label
                                    7.8
                                   GC/MS
                               analysis (Metho
                                 8270) being
                                 performed?
  7.2 Check
and adjust pH.
  7.8 Combine
  base/neutral
 extracts prior
to concentration
   7.3 -  7.6
   Extract 3
    times.
                                 7.9 -  7.11
                                Concentrate
                                  extract.
      7.7
    Further
   extractions
   required?
                                  7.12
                                Ready for
                                analysis.
                              3510B  - 8
                                                                     Revision  2
                                                                September 1994

-------
                                 METHOD 3520B

                      CONTINUOUS  LIQUID-LIQUID  EXTRACTION
1.0   SCOPE AND APPLICATION

      1.1   This method describes  a  procedure  for isolating organic compounds
from  aqueous  samples.   The  method also  describes concentration  techniques
suitable  for  preparing the  extract for  the appropriate  determinative steps
described in Sec. 4.3 of Chapter Four.

      1.2   This method  is applicable to  the  isolation and  concentration of
water-insoluble and slightly  soluble organics  in  preparation  for a variety of
chromatographic procedures.

      1.3   Method 3520 is  designed for extraction solvents  with greater density
than the  sample.   Continuous extraction devices  are available  for extraction
solvents that are less dense than the sample.  The analyst must demonstrate the
effectiveness of any  such  automatic extraction device  before employing it in
sample extraction.


2.0   SUMMARY OF METHOD

      2.1   A measured  volume of  sample,  usually  1 liter,  is   placed  into  a
continuous liquid-liquid extractor, adjusted, if necessary, to a specific  pH (see
Table 1), and extracted with  organic solvent for  18-24  hours.   The extract is
dried, concentrated (if necessary),  and,  as  necessary, exchanged  into a solvent
compatible with the cleanup or determinative method  being employed (see Table 1
for appropriate exchange solvents).


3.0   INTERFERENCES

      3.1   Refer to Method 3500.

      3.2   Under basic extraction conditions required to separate analytes for
the packed columns of Method 8250, the decomposition of some analytes has been
demonstrated.   Organochlorine pesticides  may dechlorinate, phthalate esters may
exchange, and  phenols  may react to form tannates.   These  reactions increase with
increasing pH,  and  are decreased by the shorter reaction  times  available in
Method 3510.   Methods 3520/8270, 3510/8270, and  3510/8250,  respectively,  are
preferred over Method  3520/8250 for the analysis of  these classes of compounds.


4.0   APPARATUS AND MATERIALS

      4.1   Continuous liquid-liquid extractor  -  Equipped with Teflon or glass
connecting joints and  stopcocks  requiring  no lubrication (Kontes 584200-0000,
584500-0000, 583250-0000,  or equivalent).

      4.2   Drying column  -  20 mm ID Pyrex chromatographic column  with Pyrex
glass wool at bottom and a Teflon stopcock.

                                   3520B  -  1                         Revision  2
                                                                September 1994

-------
      NOTE: Fritted glass  discs are difficult  to decontaminate  after  highly
            contaminated extracts  have  been  passed through.   Columns without
            frits may  be  purchased.   Use a  small  pad of Pyrex glass wool  to
            retain the  adsorbent.   Prewash the glass  wool  pad with  50  ml  of
            acetone followed by 50 ml of  elution  solvent  prior to packing the
            column with adsorbent.

      4.3   Kuderna-Danish (K-D) apparatus

            4.3.1 Concentrator tube -  10  ml graduated  (Kontes K-570050-1025 or
      equivalent).  A  ground  glass stopper is used to  prevent evaporation  of
      extracts.

            4.3.2 Evaporation  flask  -    500  ml  (Kontes  K-570001-500  or
      equivalent).   Attach  to  concentrator  tube  with  springs,  clamps,  or
      equivalent.

            4.3.3 Snyder column  -   Three ball macro  (Kontes  K-503000-0121  or
      equivalent).

            4.3.4 Snyder column  -    Two  ball micro  (Kontes  K-569001-0219  or
      equivalent).

            4.3.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.4   Boiling chips  - Solvent extracted,  approximately  10/40 mesh (silicon
carbide or equivalent).

      4.5   Water  bath  -  Heated,  with  concentric   ring  cover,  capable  of
temperature control (± 5°C).   The bath should  be used  in a hood.

      4.6   Vials - 2 ml,  glass with Teflon lined  screw-caps or crimp tops.

      4.7   pH indicator paper - pH range including the desired extraction pH.

      4.8   Heating mantle - Rheostat controlled.

      4.9   Syringe - 5 ml.


5.0   REAGENTS

      5.1   Reagent grade  chemicals shall  be used in all  tests. Unless otherwise
indicated, it is intended  that all  reagents shall conform  to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are available. Other grades may  be used, provided it is first
ascertained that the reagent  is of sufficiently high  purity to permit its use
without lessening the accuracy of the determination.  Reagents should  be stored
in glass to prevent the leaching of contaminants from plastic containers.

      5.2   Organic-free reagent water  - All references  to water in this method
refer to organic-free reagent water,  as defined  in Chapter One.
                                   3520B  -  2                         Revision 2
                                                                September 1994

-------
      5.3   Sodium  hydroxide solution  (ION),  NaOH.   Dissolve  40 g  NaOH  in
organic-free reagent water and dilute to 100 mL.

      5.4   Sodium sulfate (granular, anhydrous), Na2S04.   Purify by heating  at
400°C for 4 hours in a shallow tray, or by precleaning the sodium  sulfate with
methylene chloride.   If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed,  demonstrating that there is no interference from
the sodium sulfate.

      5.5   Sulfuric acid solution  (1:1 v/v),  H2S04.   Slowly  add 50 ml  of H2S04
(sp. gr. 1.84) to 50 ml of organic-free reagent water.

      5.6   Extraction/exchange solvents

            5.6.1 Methylene chloride, CH2C12 - Pesticide quality or equivalent.

            5.6.2 Hexane, C6H14  -  Pesticide quality  or equivalent.

            5.6.3 2-Propanol, (CH3)2CHOH -  Pesticide quality  or equivalent.

            5.6.4 Cyclohexane, C6H12 -  Pesticide  quality or equivalent.

            5.6.5 Acetonitrile, CH3CN - Pesticide quality or equivalent.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory material  to this  chapter,  Organic Analytes,
Sec. 4.1.


7.0   PROCEDURE

      7.1   Using a 1 liter graduated cylinder, measure out 1  liter  (nominal)  of
sample  and  transfer it quantitatively  to  the  continuous  extractor.   If high
concentrations are anticipated,  a  smaller  volume may be  used and  then diluted
with organic-free reagent water to 1 liter.  Check the  pH of the sample with wide-
range pH paper and adjust the pH, if necessary,  to the pH  indicated in Table 1
using 1:1  (V/V)  sulfuric  acid or 10 N sodium hydroxide.   Pipet 1.0 mL of the
surrogate standard spiking solution into each sample  into  the  extractor and mix
well.  (See Method 3500 and  the  determinative method  to be  used, for details  on
the surrogate standard solution and the matrix spike  solution.)  For the sample
in each analytical batch selected for spiking, add 1.0 mL  of the matrix spiking
standard.   For  base/neutral-acid analysis, the amount of the  surrogates and
matrix  spiking   compounds  added  to  the  sample  should   result   in  a  final
concentration of  100 ng//iL of each  base/neutral  analyte  and  200 ng/juL of each
acid analyte  in  the extract  to be  analyzed  (assuming a  1 /nL injection).    If
Method 3640,  Gel-Permeation  Cleanup,  is to be  used, add  twice  the  volume  of
surrogates and matrix spiking compounds since half  the extract is  lost due  to
loading of the GPC column.

      7.2   Add 300-500 mL of methylene chloride to the distilling flask.  Add
several  boiling chips to the flask.


                                   3520B  -  3                         Revision 2
                                                                September 1994

-------
      7.3   Add sufficient water to the extractor to ensure proper operation and
extract for 18-24 hours.

      7.4   Allow to cool; then detach the  boiling  flask.   If extraction at a
secondary pH is not required (see Table 1), the extract is dried and concentrated
using one of the techniques referred to in Sec. 7.7.

      7.5   Carefully,  while  stirring, adjust the pH of the aqueous phase to the
second pH  indicated  in Table 1.    Attach  a clean  distilling  flask  containing
500 ml of methylene  chloride  to  the continuous extractor.   Extract  for 18-24
hours, allow to cool, and detach  the distilling flask.

      7.6   If performing GC/MS analysis (Method 8270), the acid/neutral and base
extracts may be combined prior to  concentration.   However,  in some situations,
separate concentration  and analysis of the acid/neutral and base extracts may be
preferable (e.g.  if for regulatory  purposes  the presence or absence of specific
acid/neutral  and  base  compounds  at  low  concentrations  must be  determined,
separate extract analyses may be  warranted).

      7.7   Perform concentration  (if necessary) using the Kuderna-Danish (K-D)
Technique (Sees.  7.8.1  through 7.8.4).

      7.8   K-D Technique

            7.8.1 Assemble a Kuderna-Danish  (K-D) concentrator by attaching a 10
      ml concentrator tube to a 500  ml evaporation flask.   Dry the  extract by
      passing it through  a drying  column  containing  about  10  cm  of anhydrous
      sodium sulfate.  Collect the dried extract in a K-D concentrator.   Rinse
      the flask which contained the solvent extract with  20-30 mL  of methylene
      chloride and add  it to  the  column to complete the quantitative transfer.

            7.8.2 Add one or  two  clean boiling chips to the  flask and attach a
      three ball  Snyder column.   Prewet the Snyder column by adding  about 1 ml
      of methylene chloride to the top of the column.  Place the K-D apparatus
      on a hot water bath (15-20°C above  the  boiling  point  of the solvent)  so
      that the concentrator tube is partially  immersed  in the hot water and the
      entire  lower  rounded surface  of the  flask  is  bathed  with  hot  vapor.
      Adjust the vertical position of the apparatus and the  water temperature,
      as required,  to  complete the  concentration  in  10-20  minutes.    At  the
      proper rate of distillation the balls of the  column will actively chatter,
      but the  chambers will  not  flood.   When the apparent volume  of  liquid
      reaches 1 ml,  remove the K-D apparatus from the water bath and allow it to
      drain and  cool for at  least  10  minutes.   Remove the  Snyder  column  and
      rinse the flask and its lower joints into the concentrator tube with 1-2
      ml of extraction  solvent.

            7.8.3 If a  solvent exchange is required (as indicated  in Table 1),
      momentarily remove the  Snyder column, add 50 ml of the exchange solvent,
      a  new  boiling chip, and  reattach   the  Snyder column.   Concentrate  the
      extract, as described in Sec.  7.9,  raising  the temperature  of the water
      bath, if necessary, to  maintain proper distillation.
                                   3520B  -  4                         Revision 2
                                                                September 1994

-------
            7.8.4 Remove the  Snyder  column  and rinse the  flask  and its lower
      joints into the  concentrator  tube with 1-2 mL  of  methylene chloride or
      exchange solvent.   If sulfur  crystals are a  problem,  proceed to Method
      3660 for cleanup.  The  extract  may  be further concentrated by using the
      techniques outlined in  Sec. 7.9 or  adjusted to  10.0 ml with the solvent
      last used.

      7.9   If further concentration is  indicated in Table  1, either the micro-
Snyder column technique (7.9.1) or nitrogen blowdown technique (7.9.2) is used
to adjust the extract to the final volume required.

            7.9.1 Micro-Snyder Column Technique

                  7.9.1.1     Add another one or two clean  boiling chips to the
            concentrator tube and  attach a two ball micro-Snyder column.  Prewet
            the  column  by  adding 0.5  ml of  methylene  chloride  or  exchange
            solvent to the top of  the column.   Place the K-D apparatus in a hot
            water bath so that the  concentrator tube  is  partially immersed in
            the hot water.  Adjust  the  vertical position  of the  apparatus and
            the water temperature, as required, to complete the  concentration in
            5-10 minutes.  At the proper rate of distillation the balls of the
            column will  actively chatter,  but the chambers will  not flood.  When
            the  apparent volume  of  liquid reaches  0.5 ml,  remove  the  K-D
            apparatus from the water bath  and allow  it to drain and cool for at
            least 10 minutes.   Remove the Snyder column, rinse the flask and its
            lower joints  into the concentrator tube with  0.2  mL of methylene
            chloride or exchange  solvent,  and adjust the final  volume to 1.0 to
            2.0 ml,  as indicated in Table 1, with  solvent.

            7.9.2 Nitrogen Blowdown Technique

                  7.9.2.1     Place the  concentrator tube  in  a  warm bath (35°C)
            and evaporate the solvent volume to  0.5  ml using a gentle stream of
            clean, dry nitrogen (filtered through a column  of activated carbon).

                  CAUTION:    New plastic tubing must not  be  used between the
                              carbon  trap   and  the  sample,   since   it   may
                              introduce  interferences.

                  7.9.2.2     The internal wall  of the tube must  be rinsed down
            several  times with methylene chloride or appropriate  solvent during
            the operation.  During evaporation, the  tube  solvent  level must be
            positioned to avoid water condensation.   Under normal procedures,
            the extract must not  be allowed to become dry.

                  CAUTION:    When the volume of solvent is reduced below 1 ml,
                              semivolatile analytes  may  be lost.

      7.10  The extract may now be  analyzed  for the target analytes using the
appropriate determinative  technique(s)  (see  Sec.  4.3 of  this  Chapter).    If
analysis  of the  extract  will  not  be  performed  immediately,  stopper  the
concentrator tube and store  refrigerated.   If the extract will be stored longer
                                   3520B  -  5                         Revision 2
                                                                September 1994

-------
than 2 days it should be transferred  to  a vial with a Teflon lined screw-cap or
crimp top, and labeled appropriately.


8.0   QUALITY CONTROL

      8.1   Any reagent  blanks,  matrix  spike,  or replicate  samples  should be
subjected to  exactly the same analytical  procedures  as those  used  on actual
samples.

      8.2   Refer to  Chapter  One for specific quality  control  procedures and
Method 3500 for extraction and sample-preparation procedures.


9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods for performance data.


10.0  REFERENCES

1.    U.S. EPA 40  CFR Part 136, "Guidelines Establishing  Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act;  Final Rule and Interim
      Final Rule and Proposed Rule," October 26,  1984.
                                   3520B - 6                        Revision 2
                                                                September  1994

-------





















GO
Q
O

1—
LU


LU

i — i
1—

t— 1
21
a:
LU
h—
LU
Q

GO
^-^
o
H— (

^»
^^
•*^

rv
O
) i

GO
o

1 —

Q

o
i_~t

o
t—
o
cc
h—
X
LU

O

LU
1— 1
o
LU
Q.
GO




























(J
i— rO
rO S-
C +•*
•i- X
LU UJ






•fj
o
rO
^_
QJ •+->
E X
3 QJ
O l+-
> O










QJ
CT>+->
C C
rO QJ
J= >
O r—
X O
LU 01











QJ
a>+->
c c
rO QJ
-C >
O .—
X 0
LU 00










































QJ

3
_. __ e
r— • >-
O O
> M—







•o
QJ
s-

3
QJ O
S- lt-










T3
QJ
S-
3
o- s-
QJ O
C tt











-a
QJ
s-

3
CT S-
QJ O
S- <+-




c
D.0-
ro -(->
•o o
C 05
O S-
O +->
QJ X
GO QJ



C
0
r— +J
rO O
••- ro
+•> S-
•!- +J
C X
h-l QJ




QJ
>
•'-
rO
C
• r—
E
s-
QJ
Q
„ — ,.
_J
E
^-^
00
oo
^
c
TO





^
_i
" — '
a.
3
C
rO
U










O.
3
C
rO
a>

(_)











00
• r—
00
>>

fl3
c:
a?










"^
CL










T"
Q.










-o
0
HI
VJU


to
Q

0
i ni.^

oooooooooooo
f— i O O O O O •— i •— i O •— i •— < O












OOOOOOOOOOOO
.— (CMCMCMOOCMCMCMCMCMO



QJ
•o

s_
o
(J QJ
c
QJ rO
C X
QJ QJ
QJ QJQJi — QJQJQJ-CajQJQJQJ
CCC>>CCCOCCCC
XXX-MXXXOXXXX
QJ QJ QJ QJ QJ QJ QJ >) QJ QJ QJ QJ
f~ f— f— ^z f-~ (•— »•— (j <*•• f— (~ f—










O
C r—
ro O
Q. QJQJCQJQJQJ QJQJQJQJ
OCCfOCCC CCCC
S— rO ro r~ rOrOrOQJrOrOrorO
Q.XX-I-JXXXCXXXX
1 Q)Q)Q)QJQJQJOQJQJQJQJ
CMJ=J= E-CJC.C CJ=^IJ=J^









QJQJQJQJQJQJQJQJQJQJQJQJ
CCCCCCCCCCCC
OOOOOOOOOOOO
CCCCCCCCCCCC


-o-o-o -o -o -o -o
QJ QJ QJ QJ QJ QJ QJ
> > > > > > >
QJ Q) QJ QJ QJ QJ QJ
O U O O O O O
QJ Q) QJ QJ QJ QJ QJ
S- S- S- S- S- S- S-
CT> CT> en ra
CMOOOOOOI 1 1000001001
VlrOroromLOLOrorOroroiO












OO-— iOO^-iOOOC3r-iO
ooracoraraoorarararaoooo






o o
O r- 1












O
o
1— 1










QJ
C
X i
OJ
f»













OJ
c
ro QJ
X C
QJ O
-C C









QJ
c
O CM
C V


_,_,
QJ
>
QJ
U
QJ
i_
f—i
00 •— I
ro A










y
_Q
5S
CO CO






o o o
,-. r-.^,













1 1 1












1 1 1










QJ
•r-
s_
+-> r—
•i- o
C C
O rO
QJ +-> -C
C QJ -l->
O O QJ
C ro £









QJ QJ
•— i C C
i— i O O
A C C


•o -o
QJ QJ
> >
•r— 'r—
QJ QJ
O O
QJ QJ
S- S-

CM 00 01
V ro rO










•g
^
° 2^
^j co co
ro oo oo


"^
s_
-o

o
o












o
o



QJ
-o

S-
o
J^
o
QJ
C
QJ
i —
j=
I *
QJ
E



QJ
T3

S-
o
f—
(J
QJ
C
QJ

'X)
J=
+J
QJ










QJ
C
o
c


-o
QJ
>
QJ
O
QJ
S-

00
rO












2
*lf
OT
r— O-
rO 3
C C
O ro
-r- Q)
+-> r—
Q. 0
O
QJ
ro 5

oo _
c 0

1 §
° -S

^lS CT)
05 ^v S-
CD TJI fO
•*J ^"^
O
-° O
-a us
o "o co
c- QJ
+J Nl-o
QJ >> O
^~ r~~ r-
ro ^3
5 QJ
• rO ^~
Q "*•
i_l_ ^j

^•^ Q
Ja'tj ^
f^ ^*"
4-> i_
O -)->
rO X
S- QJ >•>
•4-" i-
X QJ ro
QJ C f>
rO 01
r~™" 3^ QJ
O QJ U
C j= QJ
rO C
0-_J C
O £ 3
S-
Q-o "1
CM (J
ro ro
-^ S-
E c 4J
_ -r- X
0 ru
• (/)
*~~1 -(-^ QJ
„, i— -C
0)^"^
— *- o
00 j_
-3 o 3
*- 'r~ c
O "£ W
^J- s OJ
O i —

^^ 1 —
-0 g QJ
^ QJ nrt ~O
T3 jz 2 QJ
QJ °- S-
s: s- >>'5
^•2 E «
QJ CO
-0s-?14-
^ -2 o *"~
^L QJ ^^ QJ
•^ /j
r~"" Q (4™ f+i
i- O rrt
*"~ (^
>!•!—
OJ C -M rO
_Q O -f- >
•r- U fO
>^ +_> ••—
t^ rO 4— (/)
E N •>- QJ
••- 0 S_
i^ rO O-"O
O > 00 Q)
C -r- (J
QJ i- QJ O
c— O) ^— £_
a. -o h- Q.
ro JO
00
• *r—
. — ,
CM ^
rr> °-
_
o "*
0)^
^-^
QJ
QJ C
00 o
00 -r-
o "st
•r- rO
"i OJ
0 £
ry^
1
T3
.° ••-
0 «*
ro
S- -
-t-> O
X LO
QJ ID
ro
QJ
£-0
QJ
"*"* "QJ
QJ S
c -
3 -0
s-2
3 -r-
(J 3
LJ ^y
O QJ
S-

rO
O /"
'f~
01
i — oo
O +->
C C
QJ QJ
0- 0
a.
"2 E
c o
ro (_>

rt
~^ ro
— S-
." =
(/) C
QJ
Q.-Q
QJ «
C
i- ^
o ""^
r-- °
J= ro
rt ^
_ _.
fO *o

°^_
" Q
'C.
QJ C
"*~ f
*
/«
t
_L_J )n
rO Q.
i — QJ
ro 01

+J j_

O_ ^
^
o i:
^}
01 **•
00
o ^+~
	 1 1—1
O T3
4_>
00
S-
•1 —
**"
QJ
rO
-°
QJ
-C
-f->
S-
1
I?
0)
>


QJ
£_
O
QJ
rO

00
QJ
oo
oo
O
1
-2
r-
C
ro
00

"O
ft)
~?
Sj
~
—•
r^
fc
O
OJ
S—
-*->
c

OO
•r—

QJ
O
c
QJ
^y
QJ
oo
^
O-
o
1 —
oo •

"O CM
o •
1 »
QJ •
y o
OJ GO
•4-> QJ
OJ
r | ^^
O , 	 .
I— c
O
^ .f*.
+^>
OJ ..
rO
rv^ S^
CC. ^j
X
. QJ
-o
QJ (/)
T£5 ^J
c o
E c

0 +->
a c
QJ O
S- U

CM •*
O1
C  -Q
QJ E
rv QJ
+->
Q.
QJ
GO






























r^

i
CO
o
CM
LD
CO















































-------
                              METHOD  3520B
               CONTINUOUS  LIQUID-LIQUID  EXTRACTION
7.1  Add appropriate
   surrogate and
   matrix spiking
    solutions.
     7.7 - 7.8
Concentrate extract
7.2 Add methylene
    chloride to
  distilling flask.
      7.8.3 Is
      solvent
     exchange
     required?
     7.8.3 Add
  exchange solvent;
concentration extract
 7.3 Add reagent
water to extractor
 extract for 18-24
      hours.
    7.9 Further
 concentrate extract
    if  necessary;
 adjust final volume.
 7.5 Adjust pH of
 aqueous phase;
 extract for 18-24
 hours with clean
      flask.
 7.10 Analyze using
 organic techniques.
       7.6
     GC/MS
     malysis
  (Method 8270)
   performed?
        8000
       Series
      Methods
 7.6 Combine acid
 and base/neutral
 extracts prior to
  concentration.
                                3520B -  8
                                     Revision 2
                                September  1994

-------
                                 METHOD 3540B

                              SOXHLET EXTRACTION
1.0   SCOPE AND APPLICATION
      1.1   Method 3540  is  a procedure  for  extracting nonvolatile  and  semi-
volatile organic compounds from  solids  such as soils, sludges, and wastes.  The
Soxhlet extraction process ensures intimate contact of the sample matrix with the
extraction solvent.

      1.2   This method is applicable to the isolation and concentration of water
insoluble and slightly water  soluble organics in preparation  for a variety of
chromatographic procedures.


2.0   SUMMARY OF METHOD

      2.1   The solid sample is  mixed with anhydrous sodium sulfate, placed in
an extraction thimble or  between two plugs of  glass wool,  and extracted using an
appropriate  solvent   in  a  Soxhlet  extractor.   The  extract  is  then  dried,
concentrated  (if necessary),  and,   as necessary,  exchanged  into  a  solvent
compatible with the cleanup or determinative  step being employed.


3.0   INTERFERENCES

      3.1   Refer to Method 3500.
4.0   APPARATUS AND MATERIALS

      4.1   Soxhlet extractor - 40 mm ID, with 500 mL round bottom flask.

      4.2   Drying column  -  20  mm ID Pyrex chromatographic  column  with Pyrex
glass wool at bottom.

      NOTE: Fritted  glass  discs are difficult  to decontaminate  after  highly
            contaminated extracts  have  been  passed through.   Columns  without
            frits may  be  purchased.   Use a  small  pad of Pyrex glass wool  to
            retain the  adsorbent.   Prewash the glass  wool  pad with 50 ml  of
            acetone followed by 50 mL of elution  solvent  prior to packing the
            column with adsorbent.

      4.3   Kuderna-Danish (K-D) apparatus

            4.3.1 Concentrator tube - 10  mL, graduated  (Kontes K-570050-1025 or
      equivalent).   A  ground glass stopper is used to  prevent evaporation  of
      extracts.

            4.3.2 Evaporation  flask  -    500  mL   (Kontes   K-570001-500  or
      equivalent).  Attach  to  concentrator   tube  with  springs,   clamps,  or
      equivalent.

                                   3540B  -  1                         Revision 2
                                                                September 1994

-------
            4.3.3 Snyder  column  - Three  ball  macro  (Kontes  K-503000-0121  or
      equivalent).
            4.3.4 Snyder  column   -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).
            4.3.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).
      4.4   Boiling chips - Solvent extracted, approximately 10/40 mesh (silicon
carbide or equivalent).
      4.5   Water  bath  - Heated,  with  concentric  ring   cover,  capable  of
temperature control (± 5°C).   The  bath  should be used in a  hood.
      4.6   Vials - Glass, 2  ml capacity,  with Teflon  lined  screw or crimp top.
      4.7   Glass or paper thimble or glass wool - Contaminant free.
      4.8   Heating mantle -  Rheostat controlled.
      4.9   Disposable glass  pasteur pipet and bulb.
      4.10  Apparatus for determining percent dry weight.
            4.10.1      Oven  - Drying.
            4.10.2      Desiccator.
            4.10.3      Crucibles - Porcelain or disposable aluminum.
      4.11  Apparatus for grinding
      4.12  Analytical balance -  0.0001 g.
5.0   REAGENTS
      5.1   Reagent grade inorganic chemicals shall be used in all tests.  Unless
otherwise  indicated,  it  is  intended  that all  reagents shall conform  to the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where  such  specifications  are available.  Other grades may be used,
provided it is first ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.
      5.2   Organic-free reagent  water.  All  references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.
      5.3   Sodium sulfate (granular,  anhydrous), Na2S04.  Purify by heating at
400°C for 4 hours in a shallow tray, or by precleaning the sodium sulfate with
methylene chloride.  If the sodium  sulfate  is precleaned with  methylene chloride,
a method blank must be analyzed, demonstrating that there is no  interference from
the sodium sulfate.
                                   3540B  -  2                         Revision 2
                                                                September 1994

-------
      5.4   Extraction solvents

            5.4.1 Soil/sediment and aqueous  sludge  samples  shall  be extracted
      using either of the following solvent systems:

                  5.4.1.1     Acetone/Hexane   (1:1)    (v/v),   CH3COCH3/C6H14.
            Pesticide quality or equivalent.

                  NOTE: This solvent system  has  lower disposal  cost and lower
                        toxicity.

                  5.4.1.2     Methylene    chloride/Acetone    (1:1    v/v),
            CH2C12/CH3COCH3. Pesticide quality or equivalent.

            5.4.2 Other samples shall  be extracted using the following:

                  5.4.2.1     Methylene chloride, CH2C12.  Pesticide quality or
            equivalent.

                  5.4.2.2     Toluene/Methanol   (10:1)   (v/v),   C6H5CH3/CH3OH.
            Pesticide quality or equivalent.

      5.5   Exchange solvents

            5.5.1 Hexane, C6H14.  Pesticide quality or equivalent.

            5.5.2 2-Propanol, (CH3)2CHOH.   Pesticide  quality or  equivalent.

            5.5.3 Cyclohexane,  C6H12.  Pesticide  quality or  equivalent.

            5.5.4 Acetonitrile, CH3CN.   Pesticide quality  or equivalent.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory material to this chapter, Organic Analysis, Sec.
4.1.


7.0   PROCEDURE

      7.1   Sample Handling

            7.1.1 Sediment/soil samples - Decant  and  discard any water layer on
      a sediment sample.  Mix sample thoroughly, especially  composited samples.
      Discard any foreign objects such  as sticks, leaves,  and rocks.

            7.1.2 Waste  samples  -  Samples consisting of  multiphases  must  be
      prepared by the phase separation method  in Chapter Two before extraction.
      This procedure is for solids only.

            7.1.3 Dry waste samples amenable  to  grinding  -  Grind  or otherwise
      subdivide the waste so that it either passes through a 1 mm sieve or can


                                  3540B -  3                         Revision 2
                                                                September 1994

-------
      be extruded through  a  1  mm hole.   Introduce  sufficient  sample into the
      grinding apparatus to yield at least 10 g after grinding.

            7.1.4 Gummy, fibrous,  or  oily materials not  amenable  to grinding
      should  be  cut,  shredded,  or  otherwise broken up  to allow  mixing,  and
      maximum exposure of the sample surfaces  for extraction.  The professional
      judgment of the  analyst is required for handling these difficult matrices.

      7.2   Determination  of  sample % dry  weight  - In certain  cases,  sample
results are desired based  on dry weight  basis.   When  such data are desired, a
portion of sample for  this  determination  should  be weighed out  at the same time
as the portion used for analytical  determination.

      WARNING:    The drying  oven   should  be  contained in  a   hood  or  vented.
                  Significant laboratory  contamination may result from a heavily
                  contaminated hazardous waste sample.

      However, samples known  or suspected to contain significant concentrations
of toxic, flammable,  or explosive constituents should not be oven dried because
of concerns for personal safety.  Laboratory discretion is advised.  It may be
prudent to delay oven drying of the weighed-out portion until  other analytical
results are available.

            7.2.1 Immediately after weighing the sample for extraction, weigh 5-
      10 g of the sample into a tared crucible.   Determine the % dry weight of
      the sample by drying overnight  at  105°C.   Allow  to  cool  in a desiccator
      before weighing:

            % dry weight = q of dry sample x  100
                             g of sample

      7.3   Blend 10 g of the solid  sample with 10 g  of anhydrous sodium sulfate
and place in an extraction thimble.   The extraction thimble must  drain freely for
the duration  of the extraction  period.   A glass wool  plug above and below the
sample in the Soxhlet extractor  is  an acceptable alternative  for the thimble.
Add 1.0 ml of the surrogate standard spiking solution onto the sample (see Method
3500 for details on the surrogate standard and matrix spiking solutions).  For
the sample  in each  analytical batch  selected for  spiking, add  1.0 ml  of the
matrix spiking standard.  For base/neutral-acid analysis, the amount added  of the
surrogates and matrix spiking compounds  should result in a final concentration
of 100 ng/juL of each base/neutral analyte and  200 ng//iL of each acid analyte in
the extract to be analyzed (assuming  a  1 /xL  injection).   If  Method 3640, Gel
Permeation  Chromatography  Cleanup, is  to  be used, add  twice the  volume  of
surrogates and matrix spiking  compounds  since half  the  extract is lost due to
loading of the GPC column.

      7.4   Place approximately 300 mL of the extraction solvent (Sec. 5.4) into
a 500 ml round bottom flask containing one or two clean boiling chips.   Attach
the flask  to  the  extractor  and  extract the sample  for  16-24 hours  at 4-6
cycles/hr.

      7.5   Allow the extract to cool  after the extraction is complete.
                                   3540B -  4                         Revision 2
                                                                September 1994

-------
      7.6   Assemble  a Kuderna-Danish  (K-D)  concentrator  (if  necessary)  by
attaching a 10 ml concentrator tube to a 500 ml evaporation flask.

      7.7   Dry the  extract  by passing it through  a  drying column containing
about 10 cm of anhydrous  sodium sulfate.   Collect  the dried  extract  in a K-D
concentrator.   Wash the extractor flask and  sodium sulfate column with 100 to 125
ml of extraction solvent to complete the quantitative transfer.

      7.8   Add one or two clean boiling chips to the flask and attach a three
ball Snyder column.   Prewet the Snyder column  by adding about  1 mL of methylene
chloride to the top  of the column.   Place the  K-D apparatus  on a hot water bath
(15-20°C above the boiling point of the solvent)  so that the concentrator tube
is partially immersed in the hot water and the entire lower rounded surface of
the  flask  is   bathed  with hot  vapor.  Adjust  the  vertical  position  of  the
apparatus and  the water temperature,  as required, to complete the concentration
in 10-20 minutes.  At  the  proper  rate of distillation the balls of the column
will actively  chatter,  but the chambers will not flood.  When  the apparent volume
of liquid reaches 1-2 mL,  remove the K-D apparatus  from the water bath and allow
it to drain and cool  for at least 10 minutes.

      7.9   If  a solvent  exchange  is  required  (as  indicated  in Table  1),
momentarily remove the Snyder  column,  add  approximately  50  ml of the  exchange
solvent and a  new boiling chip, and reattach the Snyder column.  Concentrate the
extract as described  in Sec.  7.8,  raising the  temperature  of the water bath, if
necessary, to  maintain proper distillation.   When the apparent  volume again
reaches 1-2 mL, remove the K-D  apparatus from the  water  batch and allow it to
drain and cool for at least 10 minutes.

      7.10   Remove the Snyder column and  rinse the flask and its lower joints
into the concentrator tube with 1-2 mL of methylene chloride or exchange solvent.
If sulfur  crystals  are a  problem,  proceed to Method 3660  for cleanup.   The
extract may be further concentrated  by using  the  techniques described  in  Sec.
7.11 or adjusted to 10.0 mL with the solvent last  used.

      7.11  If further  concentration is indicated in Table 1,  either micro Snyder
column technique (Sec.  7.11.1) or nitrogen  blowdown technique (Sec. 7.11.2) is
used to adjust the extract to the final volume required.

            7.11.1      Micro Snyder Column Technique

                  7.11.1.1    Add another one  or two clean boiling chips to the
            concentrator tube and  attach a two  ball  micro Snyder column.  Prewet
            the column  by  adding about  0.5 mL  of methylene chloride or exchange
            solvent  to  the top of the  column.  Place the K-D apparatus in a hot
            water bath so that the concentrator tube  is  partially immersed in
            the hot water.  Adjust the vertical position  of the apparatus  and
            the water temperature, as required, to complete the concentration in
            5-10 minutes.   At the proper  rate  of  distillation the balls of the
            column will  actively chatter, but the  chambers  will not flood.  When
            the  apparent  volume  of liquid  reaches  0.5  mL,  remove  the  K-D
            apparatus from the water  bath and  allow  it to drain and cool for at
            least 10  minutes.   Remove  the Snyder column and rinse the flask and
            its  lower  joints with  about  0.2  mL  of  solvent and  add   to  the


                                  3540B - 5                         Revision  2
                                                                September  1994

-------
            concentrator tube.   Adjust  the final  volume to  1.0-2.0 ml,  as
            indicated in Table 1, with solvent.

            7.11.2      Nitrogen Slowdown Technique

                  7.11.2.1     Place the concentrator tube in  a warm water bath
            (approximately   35°C)  and  evaporate  the  solvent  volume  to  the
            required  level   using  a  gentle  stream  of  clean,  dry  nitrogen
            (filtered through a column of activated carbon).

                  CAUTION:     Do not use plasticized tubing between the carbon
                              trap and the sample.

                  7.11.2.2     The internal wall  of the  tube must be rinsed down
            several  times with  the  appropriate  solvent  during  the operation.
            During evaporation,  the solvent level in the tube must be positioned
            to prevent water from condensing  into  the sample (i.e., the solvent
            level should be below the  level  of  the water bath).   Under normal
            operating conditions, the  extract should  not be  allowed to become
            dry.

                  CAUTION:     When the volume of solvent  is reduced below 1 ml,
                              semi volatile analytes may be lost.

      7.12  The extracts obtained may  now be analyzed  for the target analytes
using the appropriate organic technique(s) (see  Sec. 4.3  of this Chapter).  If
analysis  of  the extract  will  not  be  performed  immediately,  stopper  the
concentrator tube and store  in  a  refrigerator.   If the extract will be stored
longer than 2  days,  it should be transferred to a vial  with a Teflon lined screw
cap or crimp top, and labeled appropriately.


8.0   QUALITY CONTROL

      8.1   Any reagent blanks  or matrix  spike  samples should be subjected to
exactly the same analytical  procedures as those  used on  actual samples.

      8.2   Refer to  Chapter One for  specific  quality control  procedures and
Method 3500 for extraction and sample preparation procedures.


9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods for performance data.
10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final Rule and Interim
      Final Rule and  Proposed Rule," October 26, 1984.
                                   3540B -  6                        Revision 2
                                                                September 1994

-------


1
1













oo
t~~\
O
3:
h-
LU
S;

1 1 i
>
HH
|—
-H
SI
o;
LU
I—
LU
Q

OO
o
t— (
o:

• >
i — <
o:
LU O
	 1 U_
CQ
eC OO
1— Z
O
HH
h-
>— i
O
"Z.
o
o

z
o
t— H
I —
X
UJ

f)
t— 1
LJ_
1—4
0
LU
Q_
oo































u_







QJ
E
3
O












CU
CT
£
(TJ
-C
o
X
LU










CU
CT
£
re
u
X
LU




































_,_,
U
S_
X
0)




4J
u
ro
i-
X
CU

4-
O










4->
C
cu
>
r—
O
CO











4->
£
cu
>
'o
CO




































CU
E i-
3(^
(J
i— 4-







-o
CU
S-
3 O
w 4—
CU











•o
CU
s-
•i- S-
3 O
CT4-
QJ
S_










•o
cu
S-
•1- i-
3 O
CU




C
o
.J3
(J
r
A—
X
LU



CU
;>
•r—
| *
ro
£

E
CU
4-J
cu
Q









1
-•J
E

CO

>>
're
C
rO




^7
E
*—^

3
£1
re
cu
r—
U









CL
3
C
re
cu










CO
.r—
CO
"re
c







o.








H£3
o

+->
E












"o *—•
o s-
r-H T3

cToOOOOOOOOOOOO O OOO
i — 1 O O O O O i-H i-H O »™l »-H O O »"H i-H f-H i — 1 O









O 0 O o O 0 O 0 O O 0 0 0 o
i— ICVJCVJpJOOCVJCVJCVJCVJCVJOO 1 1 1 IQ
I 	 1 r-H 1 — 1 r-H , 	 |




cu cu
^O t3
-r— «i^
s_ s.
0 0
i — r—
u cu u
c
cu re cu
ex c
cu cu cu
CUQJCUr— CUQJQJ-CQJCUQJQJQJ i—
CCC^CCCOCCCCC >>
rorerejZdjrerOi^rorererorOi i i i -C
XXX4->XXXOXXXXXi i I i4->
CUCUCUCUCUCUCU>>CUCUCUCUCU CU
JCJ=J=EJ=J=J=OJ=J=J=J=J= E

cu
•o

s_
o

cu j=
i— O

'o si cu
C r— 4-> i— C
re o •!- o cu
Q. QJ QJ C: OJ QJ QJ QJ QJ QJ QJ QJ CCi—
occraccc ccccc o re >,
s- rere-crererecurerererereoj o>4->jcjc
ca.xx4-»xxxcxxxxxc:ccu4->4->
i CUCUCUCUCUCUOCUCUCUCUCUOOOCUCU
cvi -C jz: E -c: .£= -C CJCJCJCJCTJC: c: c re E E



_^ 113} ^j (DCl)(DCl)(l)(D CD CD Q) CD
>>>>>>>>>>>>>>>>>>
CDCDtDCDaJCDtDaJCDCDCDClJCDCD (/)C/)«/>(/>(/) CO C/)(/JCO
fO rO rO (13 rtf n3 (Q fO (& (& fQ (T3 ("0 (0 H3 f0 ^9 ("0







u u
n a (o
OOi— lOOr- (OOOOi— iO>— (OOO.— (O
^- vo^or^ooooo^O'— ICSJCNJ^I-^-LO r-^. •— • csj *-H
000000000000000000000000000000000000



c
o
• ^
4_>
(TJ
s-

c
cu
o
c
0
u
a>
c
•r™
O
r—
O
4™

-o
CU
o

s_
cu
a.
cu


-a

3
O
CO

O
in
^o
CO

-a
o
.c
+J
CD
s:
co
^^
u
re
s»-
4-*
X
0)


re
S-
^~^
3
CU
c
cu
CO
re
jQ

-a
c
re

T3
u
re
cu
4->
re
Q.
cu
CO

c
•r*
re
4J
JO
o
o
t—

a






























































	 1
E
Q
O
i-H
O
+J,
0
re
s-
4J>
X
cu

cu

+•>
4-
O


O CU
•* jn
o
00 O
4_>
•a
0 4->
-C U
4-> re
cu s-
X
cu
Q cu
I-H ^-
xT rt
o co
CD f
^»_J
c
to — ^
»r- ^
CO I"H
ro c



c jif


^j 00
o p
rtJ *-
^
-*-* (J
el)
-^
Q
C
S "o
c
o _.
C"
Q- f^
1
^ s-
— I ,°
E '4~
O CU
, ^_
•— ( 3

re cu
o
C7) o
C £_
•r- Q.
CO
3 C
O -r—
o "re
00 is|
"° +J

4-^ .^
CU j_
s cu
>^ ^-*
JD (_
T3 re
M "-
f °s
... ™ o
_Q CO O
£

re ca jo
E +->
c-o
CO O CU
i— O M
0 >>
C 0 i—
CD 00 CD
j= i— c:
o. re re

s>
CU
U
£
re
•o

3


S-
0


o
o
IO
CO

o
5
cu
IE

0
^
m
U'
cu
a:


•
^
re
CO
CO
cu
Q
CU
c
c:


CO

(_j

^

X
cu
cu
f~
. *

i.
o


2~

re
cu

u
cu
re


^)
re
E

oo
^
o
0
40
•r-
(J
•r-
4-
•^
U
CU
CL
CO
t*l t
UJ
^:
I—

u














































"
^D
CD
i-
•^
2
CT"
CU

(4-
"^
CD
^)
rO

rO
^
rt3
to
CD
3
-o
a>
o
0
s-
CL
Q-
3
£
re
cu
f—
o

cu

-I-1
£
o


CVJ

C""
o
•r—
CO
• r-
^
0)
o:





























JOl^^



CQ
O
CO





































-------
             METHOD 3540B
         SOXHLET  EXTRACTION

'
7.1
Use appropriate
sample handling
technique

'
7.2
Determine sample %
dry weight
\
t
7.3
Add appropriate
surrogate and matrix
spiking standards
i

7.4
Add extraction
solvent to flask:
extract for 16-24
hours
i

75
Cool extract



      7.6
  Assemble K-D
  concentrator
       7.7
  Dry and collect
  extract in K-D
   concentrator
       7.8
 Concentrate using
   Snyder column
 and K-D apparatus
       7.9
    Is solvent
exchange required?
                          7.12
                      Analyze using
                    organic techniques
 Proceed
to Method
 3660 for
 cleanup
 8000
 Series
Methods
       7.9
  Add exchange
     solvent,
reconcentrate extract
                3540B  - 8
                          Revision  2
                     September  1994

-------
                                  METHOD  3541

                         AUTOMATED SOXHLET EXTRACTION
1.0   SCOPE AND APPLICATION

      1.1   Method 3541 describes the extraction  of organic analytes from soil,
sediment, sludges, and waste solids.   The method  uses a commercially available,
unique, three stage extraction system to achieve  analyte recovery comparable to
Method 3540, but in a  much shorter time. There are two differences between this
extraction method  and Method  3540.   In the initial  extraction stage of Method
3541, the sample-loaded extraction thimble is immersed into the boiling solvent.
This ensures very  rapid  intimate  contact  between the specimen and solvent and
rapid extraction of the  organic  analytes.   In the  second  stage the thimble is
elevated above the  solvent, and  is rinse-extracted as  in  Method 3540.   In the
third stage, the  solvent is evaporated, as would occur  in  the Kuderna-Danish
(K-D) concentration step in  Method 3540.  The concentrated  extract  is then ready
for cleanup (Method 3600) followed by measurement of the organic analytes.

      1.2   The method is applicable to the extraction and concentration of water
insoluble  or  slightly  water  soluble polychlorinated   biphenyls  (PCBs)  in
preparation for gas chromatographic  determination using either Method  8080 or
8081.  This method is applicable to  soils, clays, solid  wastes and sediments
containing from 1 to 50 jug  of PCBs (measured  as Arochlors) per gram of sample.
It has been statistically evaluated at 5 and 50 /ng/g of Arochlors 1254 and 1260,
and  found to  be  equivalent  to  Method 3540 (Soxhlet  Extraction).    Higher
concentrations of PCBs are measured following volumetric dilution with hexane.

      1.3   The method  is also  applicable  the extraction  and concentration of
semivolatile organics  in preparation for  GC/MS analysis  by  Method  8270  or by
analysis using specific GC or HPLC methods.


2.0   SUMMARY OF METHOD

      2.1   PCBs:  Moist solid samples (e.g., soil/sediment samples) may be air-
dried and ground prior to extraction or chemically  dried with anhydrous sodium
sulfate.   The prepared sample is extracted  using 1:1 (v/v)  acetone:hexane in the
automated  Soxhlet  following the same  procedure  as  outlined  for  semivolatile
organics in Sec. 2.1.   The extract  is  then concentrated and exchanged into pure
hexane prior to final  gas chromatographic  PCB measurement.

      2.2   Other semivolatile organics: A 10-g solid sample (the sample is pre-
mixed with anhydrous   sodium  sulfate  for  certain  matrices)  is  placed  in  an
extraction thimble and usually extracted with  50 ml of 1:1 (v/v) acetone/hexane
for 60 minutes  in  the  boiling extraction  solvent.  The  thimble with sample is
then raised into the rinse position and extracted for an additional 60 minutes.
Following the extraction steps,  the extraction solvent is  concentrated to 1 to
2 ml.
                                   3541 - 1                         Revision 0
                                                                September 1994

-------
3.0   INTERFERENCES

      3.1   Refer to Method 3500.

      3.2   The extraction thimble and the o-rings used to seal the extraction
cup are both a source of  interference.   Both  should  be checked by including a
method blank and  following the extraction procedure as written.  Solvent rinsing
or  extraction,  prior  to  use,   may   be necessary   to  eliminate  or  reduce
interferences.   Viton  seals  contributed least  to the  interference  problem,
however,  even  they  contributed  some   interference peaks when  the  extraction
solvent was analyzed  by the electron  capture detector.  Use  of butyl  or EPDM
rings  are not recommended  since  they were  found to  contribute significant
background when  the  extraction  solvent was  1:1 v/v  hexane/acetone  or  1:1 v/v
methylene chloride/acetone.


4.0   APPARATUS AND MATERIALS

      4.1   Automated Soxhlet Extraction System - with  temperature-controlled oil
bath (Soxtec, or equivalent).  Tecator  bath oil (catalog number  1000-1886) should
be used with the  Soxtec.  Silicone oil  must  not be  used because it destroys the
rubber parts.  See Figure 1.  The apparatus is used in a hood.

      4.2   Accessories and  consumables for  the automated Soxhlet  system.  (The
catalog numbers  are Fisher Scientific  based  on the  use of  the  Soxtec HT-6,
however, other sources that are equivalent are acceptable.)

            4.2.1   Cellulose  extraction  thimbles  -  26  mm   ID   x   60  mm
      contamination free, catalog number 1522-0034, or equivalent.

            4.2.2   Glass  extraction cups (80 ml)  -  (set of  six required for the
      HT-6), catalog number 1000-1820.

            4.2.3   Thimble  adapters  -   (set  of  six   required for  the  HT-6),
      catalog number 1000-1466.

            4.2.4   Viton  seals  -  catalog number  1000-2516.

      4.3   Syringes -  100 and 1000 juL and  5  mL.

      4.4   Apparatus for Determining Percent Dry Weight

            4.4.1   Drying Oven.

            4.4.2   Desiccator.

            4.4.3   Crucibles, porcelain.

            4.4.4   Balance, analytical.

      4.5   Apparatus for grinding - Fisher  Cyclotec,  Fisher Scientific catalog
number 1093, or equivalent.
                                   3541 - 2                         Revision 0
                                                                September 1994

-------
      4.6   Spatula

      4.7   Graduated cylinder - 100 ml.

      4.8   Aluminum weighing dish -  VWR Scientific catalog number 25433-008 or
equivalent.

      4.9   Graduated, conical-bottom glass  tubes - 15 ml,  Kimble catalog number
45166 or equivalent, or 10 ml KD concentrator tube.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform  to the specifications
of the Committee on Analytical Reagents  of the American  Chemical Society, where
such specifications are available. Other grades may be used, provided it  is first
ascertained that the reagent  is  of sufficiently high  purity to permit its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All  references  to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Sodium sulfate (granular, anhydrous),  Na2S04.  Purify by heating at
400°C for 4 hours in a shallow tray,  or by precleaning the sodium sulfate with
methylene chloride.  A method blank must be analyzed,  demonstrating that there
is no interference from the sodium sulfate.

      5.4   Extraction solvents:

            5.4.1  Organochlorine  pesticides/PCB extraction:

                   5.4.1.1    Acetone/hexane    (1:1     v/v),    CH3COCH3/C6H14.
            Pesticide quality or equivalent.

            5.4.2  Semivolatile  organics extraction:

                   5.4.2.1    Acetone/hexane    (1:1     v/v),    CH3COCH3/C6H14.
            Pesticide quality or equivalent.

                   5.4.2.2    Acetone/methylene    chloride    (1:1    v/v),
            CH3COCH3/CH2C12.   Pesticide quality or equivalent.

      5.5   Hexane, C6H14.  Pesticide quality or equivalent.


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See the  introductory material  to this chapter,  Organic Analytes,
Sec. 4.1.
                                   3541 - 3                         Revision 0
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Sample handling

            7.1.1  Sediment/soil  samples  -  Decant  and discard any water layer
      on  a sediment  sample.    Mix  sample  thoroughly,  especially  composited
      samples.  Discard any foreign objects such as sticks, leaves,  and rocks.

                   7.1.1.1    PCBs or high-boiling organochlorine pesticides -
            Air-dry the sample at room temperature  for 48 hours in a glass tray
            or on hexane-cleaned aluminum  foil, or dry the sample  by mixing with
            anhydrous  sodium  sulfate until  a free-flowing  powder  is  obtained
            (see Sec. 7.2).

                   NOTE:      Dry, finely ground soil/sediment allows the best
                              extraction efficiency for non-volatile, non-polar
                              organics, e.g., PCBs, 4,4'-DDT, etc.  Air-drying
                              is not  appropriate for  the analysis  of  the more
                              volatile  organochlorine  pesticides  (e.g.  the
                              BHCs)  or  the  more volatile  of  the semivolatile
                              organics  because  of  losses   during  the  drying
                              process.

            7.1.2  Dried  sediment/soil  and  dry   waste  samples  amenable  to
      grinding - Grind or otherwise subdivide the waste so that it either passes
      through a 1 mm  sieve or can  be extruded through a  1  mm hole.   Introduce
      sufficient sample into the grinding  apparatus to yield at least 20 g after
      grinding.      Disassemble  grinder   between   samples,    according   to
      manufacturer's  instructions, and  clean with  soap and water,  followed by
      acetone and hexane rinses.

            NOTE:  The same warning on loss  of volatile analytes  applies to the
                   grinding process.   Grinding  should only  be  performed when
                   analyzing for non-volatile organics.

            7.1.3  Gummy,  fibrous, or oily  materials  not amenable to grinding
      should  be  cut, shredded,  or otherwise broken  up  to allow mixing,  and
      maximum exposure of the sample  surfaces for  extraction.   If grinding of
      these materials is preferred, the addition and mixing of anhydrous sodium
      sulfate  with  the  sample   (1:1)  may improve  grinding  efficiency.   The
      professional  judgment  of the  analyst is  required   for   handling  such
      difficult matrices.

            7.1.4  Multiple phase waste samples - Samples consisting of multiple
      phases  must  be prepared  by  the phase separation  method in  Chapter Two
      before extraction.  This  procedure  is for  solids only.

      7.2   For sediment/soil  (especially gummy  clay) that  is moist and cannot
be air-dried because of loss of  volatile  analytes - Mix  5 g of sample with 5 g
of anhydrous sodium sulfate in a small beaker using a spatula.   Use this approach
for any solid sample  that  requires dispersion of the sample particles to ensure
greater solvent contact throughout the sample mass.
                                   3541 - 4                         Revision 0
                                                                September 1994

-------
      7.3   Determination of sample percent dry weight  - In certain cases, sample
results are desired  based  on  dry weight  basis.   When  such data are desired, a
portion of sample for this determination  should be  weighed out  at  the same time
as the portion used  for analytical determination.

      WARNING:     The  drying oven should  be  contained  in a  hood or vented.
                   Significant  laboratory  contamination  may   result  from the
                   drying  of  a  heavily contaminated hazardous  waste  sample.

            7.3.1  Immediately  after weighing  the  sample  for extraction, weigh
      5-10 g of the sample  into  a  tared crucible.  Determine the % dry weight of
      the sample  by  drying overnight  at  105°C.   Allow to cool  in a desiccator
      before weighing:

            % dry weight = g of dry sample x 100
                              g of sample

      7.4   Check the heating  oil  level in the automated Soxhlet unit and add oil
if needed.  See service manual for details.  Set  the temperature on the service
unit at 140°C  when using hexane-acetone  (1:1,  v/v)  as  the extraction solvent.

      7.5   Press the "MAINS" button;  observe  that  the switch lamp is now "ON".

      7.6   Open the cold water tap for the  reflux  condensers.  Adjust the flow
to 2 L/min to prevent solvent loss through the condensers.

      7.7   Weigh 10 g  of  sample  into extraction thimbles.  For  samples mixed
with anhydrous sodium sulfate, transfer the  entire  contents of  the beaker (Sec.
7.2)  to  the  thimble.   Add  surrogate spikes  to  each  sample  and  the matrix
spike/matrix spike duplicate to the selected sample.

      NOTE: When  surrogate  spikes and/or  matrix  spikes  contain  relatively
            volatile compounds (e.g., trichlorobenzenes, BHCs, etc.),  steps 7.8,
            7.9, and  7.10 must be performed  quickly to avoid evaporation losses
            of these compounds.   As  the  spike is  added  to the sample in each
            thimble,   the   thimble should immediately be  transferred to  the
            condenser and  lowered  into the extraction solvent.

      7.8   Immediately transfer the thimbles containing the weighed samples into
the condensers.  Raise the knob  to the "BOILING"  position.   The magnet will now
fasten to the thimble.  Lower the knob to the "RINSING" position.  The thimble
will now hang just below the condenser valve.

      7.9   Insert the extraction cups containing boiling chips,  and load each
with 50 mL of  extraction  solvent  (normally  1:1  (v/v)  hexane:acetone,  see Sec.
5.4).  Using the cup  holder,  lower the locking  handle, ensuring that the safety
catch engages.  The cups are now clamped into position.  (The seals must be pre-
rinsed or pre-extracted with extraction  solvent prior to initial  use.)

      7.10  Move the  extraction knobs to  the "BOILING" position.  The thimbles
are now immersed  in solvent.  Set the timer for 60 minutes.   The  condenser valves
must be in the "OPEN" position.    Extract  for the preset time.
                                   3541 - 5                         Revision 0
                                                                September 1994

-------
      7.11  Move the extraction knobs to the "RINSING" position.  The thimbles
will now hang above the solvent surface.  Set timer for 60 minutes.  Condenser
valves are still open.  Extract for the preset time.

      7.12  After rinse time has elapsed, close the  condenser valves by turning
each a quarter-turn, clockwise.

      7.13  When all  but  2 to  5 ml  of solvent  have  been  collected,  open the
system and remove the cups.

      7.14  Transfer the contents of the cups to  15 ml graduated, conical-bottom
glass tubes.  Rinse the cups using hexane (methylene chloride if 1:1 methylene
chloride-acetone was used for extraction and analysis is by GC/MS) and add the
rinsates to the glass  tubes.  Concentrate the extracts to 1  to  10 ml.  The final
volume  is  dependent on  the determinative method  and the  quantitation  limit
required.  Transfer a portion to a GC vial and store  at 4°C until  analyses are
performed.

      NOTE:         The  recovery  solvent  volume  can  be  adjusted by  adding
                    solvent  at  the top  of the condensers.   For  more  details
                    concerning  use  of the extractor,  see the  operating manual
                    for  the  automated extraction system.

      7.15          Shutdown

            7.15.1      Turn "OFF" main switch.

            7.15.2      Turn "OFF" cold water tap.

            7.15.3      Ensure that all condensers are free of  solvent.  Empty
      the solvent that is recovered in  the evaporation step  into an appropriate
      storage container.

      7.16  The extract is now ready for cleanup or analysis,  depending on the
extent of interfering co-extractives.  See Method 3600 for guidance on cleanup
methods and Method 8000 for guidance  on  determinative  methods.   Certain cleanup
and/or determinative  methods may require a solvent exchange  prior to cleanup
and/or determination.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for general quality control procedures and to
Method 3500 for specific extraction and sample preparation QC procedures.

      8.2   Before processing any samples,  the analyst should demonstrate through
the analysis of an  organic-free solid matrix (e.g., reagent sand)  method blank
that  all  glassware and  reagents  are interference-free.   Each time a  set of
samples  is  extracted,  or when  there is  a  change in  reagents,  a  method blank
should be processed as a safeguard against chronic laboratory contamination.  The
blank samples should be  carried  through all stages of the  sample  preparation and
measurement.    This is  especially important because of  the  possibility of
interferences being extracted from the extraction cup seal.


                                   3541 - 6                         Revision 0
                                                                September  1994

-------
      8.3   Standard quality assurance practices should be used with this method.
Field duplicates should be collected to validate the precision of the sampling
technique.   Each  analysis  batch of 20 or less  samples must  contain:  a method
blank,  either a  matrix  spike/matrix spike  duplicate  or  a  matrix  spike and
duplicate  sample  analysis,  and  a  laboratory control   sample,  unless  the
determinative  method  provides  other  guidance.   Also,  routinely check the
integrity of the instrument  seals.

      8.4   Surrogate standards must be added to all samples when specified in
the appropriate determinative method.


9.0   METHOD PERFORMANCE

      9.1   Multi-laboratory accuracy and precision  data were obtained for PCBs
in soil.  Eight laboratories spiked Arochlors  1254 and  1260 into three portions
of 10 g of Fuller's  Earth  on three non-consecutive  days followed by immediate
extraction using Method 3541.  Six of the laboratories spiked each Arochlor at
5 and 50 mg/kg and two laboratories  spiked  each Arochlor  at  50 and 500 mg/kg.
All extracts  were  analyzed by Oak Ridge National Laboratory, Oak Ridge, TN, using
Method 8081.   These  data are listed  in  a table  found in Method 8081,  and were
taken from Reference 1.

      9.2   Single-laboratory  accuracy  data  were   obtained  for  chlorinated
hydrocarbons, nitroaromatics, haloethers, and organochlorine pesticides in a clay
soil.  The spiking concentrations  ranged  from 500 to 5000 jixg/kg,  depending on
the sensitivity of the analyte  to  the  electron  capture detector.   The spiking
solution was  mixed into the soil  during addition and then immediately transferred
to the  extraction device  and  immersed  in the  extraction solvent.   The data
represents a single  determination.   Analysis  was by  capillary column gas
chromatography/electron  capture  detector  following  Methods  8081  for  the
organochlorine  pesticides,   8091   for  the  nitroaromatics,  8111   for  the
hydrocarbons, and 8121 for  the chlorinated hydrocarbons.  These data are listed
in a table located in their respective methods and were  taken from Reference 2.

      9.3   Single-laboratory  accuracy  and  precision  data were  obtained for
semivolatile organics in  soil by spiking  at  a  concentration of 6 mg/kg for each
compound.  The spiking  solution was mixed into  the soil  during addition and then
allowed  to  equilibrate for  approximately 1  hr prior  to extraction.   Three
determinations  were   performed  and  each   extract  was  analyzed   by  gas
chromatography/mass spectrometry following Method 8270.  The low recovery of the
more  volatile compounds   is  probably  due  to   volatilization  losses  during
equilibration. These data  are listed in a Table  located in  Method 8270 and were
taken from Reference 2.
10.0  REFERENCES

1.    Stewart,  J.    "Intra-Laboratory  Recovery  Data for  the  PCB  Extraction
      Procedure"; Oak  Ridge National  Laboratory,  Oak Ridge,  TN,  37831-6138;
      October 1989.
                                   3541 - 7                         Revision 0
                                                                September 1994

-------
2.     Lopez-Avila, V. (Beckert, W., Project Officer), "Development of a Soxtec
      Extraction  Procedure  for  Extracting  Organic  Compounds  from  Soils  and
      Sediments", EPA 600/X-91/140,  US EPA, Environmental  Monitoring Systems
      Laboratory-Las Vegas,  October 1991.
                                   3541 - 8                         Revision 0
                                                                September 1994

-------
             Figure 1
Automated Soxhlet  Extraction System
          Condenser
            Thimble


        Glass Wool Plug

            Sample


    Aluminum beaker (cup)
           Hot plate
             3541 - 9
    Revision 0
September 1994

-------
                              METHOD 3541
                     AUTOMATED  SOXHLET EXTRACTION
         i
       7.1
 Use appropriate
 sample handling
    technique.
       7.2
 Add anhydrous
    Na2SO4if
    necessary
       7.3
Determine percent
   dry weight.
    7.4
 Check oil
  level in
Soxhlet unit.
       7.5
  Press "Mains"
     button.
        I
       7.6
 Open Cold water
 tap. Adjust flow.
       7.7
Weigh sample into
extraction thimbles.
  Add surrogate
      spike.
                                  7.8
                            Transfer samples
                            into condensers.
                            Adjust position of
                           magnet and thimble.
                                   1
                                   7.9
                             Insert extraction
                              cups and toad
                              with solvent.
                                   I
                                  7.10
                             Move extraction
                                knobs to
                               "Boiling" for
                                60 mins.
                                 ©
                                 3541 - 10
                                                            7.11
                                                       Move extraction
                                                          knobs to
                                                         "Rinsing" for
                                                          60 mins.
                                   7.12
                                   Close
                             condenser valves.
                                                            7.13
                                                        Remove cups.
                                    I
                                   7.14
                             Transfer contents
                                to collection
                               vials, dilute or
                               concentrate to
                                  volume.
                                    I
                                   7.15
                                 Shutdown
                                         Revision  0
                                     September  1994

-------
                                 METHOD 3550A

                             ULTRASONIC  EXTRACTION


See DISCLAIMER-1.  See manufacturer's specifications for operational settings.


1.0   SCOPE AND APPLICATION

      1.1   Method 3550  is a  procedure  for extracting nonvolatile  and semi-
volatile organic compounds from solids  such  as  soils, sludges, and wastes.  The
ultrasonic  process  ensures  intimate  contact  of  the  sample  matrix with  the
extraction solvent.

      1.2   The  method  is divided  into two sections,  based  on  the expected
concentration  of organics  in the  sample.    The   low  concentration  method
(individual organic components of < 20  mg/kg)  uses  a larger sample size and a
more rigorous extraction procedure (lower concentrations are more difficult to
extract).  The medium/high concentration method (individual organic components
of > 20 mg/kg) is much simpler and therefore faster.

      1.3   It is highly  recommended that the  extracts  be  cleaned up prior to
analysis.  See Chapter Four (Cleanup),  Sec.  4.2.2,  for applicable methods.


2.0   SUMMARY OF METHOD

      2.1   Low  concentration  method -  A 30 g  sample  is mixed  with anhydrous
sodium sulfate to form a free-flowing powder.  This is solvent extracted three
times using ultrasonic extraction.   The  extract is separated from the sample by
vacuum filtration or centrifugation.  The extract  is ready for  cleanup and/or
analysis following concentration.

      2.2   Medium/high  concentration  method  -  A  2 g sample  is  mixed  with
anhydrous  sodium sulfate  to  form  a  free-flowing  powder.   This  is  solvent
extracted once using ultrasonic extraction.  A portion of the extract  is removed
for cleanup and/or analysis.


3.0   INTERFERENCES

      3.1   Refer to Method 3500.


4.0   APPARATUS AND MATERIALS

      4.1   Apparatus  for grinding dry  waste samples.

      4.2   Ultrasonic preparation  - A horn type device  equipped with a titanium
tip, or a device that  will give equivalent performance, shall  be used.
                                  3550A  -  1                         Revision 1
                                                                September 1994

-------
            4.2.1 Ultrasonic Disrupter - The disrupter must have a minimum power
      wattage of  300 watts, with  pulsing capability.   A device  designed  to
      reduce the  cavitation sound is  recommended.    Follow  the  manufacturers
      instructions for preparing the  disrupter  for  extraction  of samples with
      low and medium/high concentration.

            Use a 3/4" horn for the low concentration  method and a 1/8" tapered
      microtip attached to  a 1/2" horn for  the medium/high concentration method.

      4.3   Sonabox - Recommended with above disrupters for decreasing cavitation
sound (Heat Systems - Ultrasonics,  Inc.,  Model  432B or equivalent).

      4.4   Apparatus for determining percent dry weight.

            4.4.1 Oven - Drying.

            4.4.2 Desiccator.

            4.4.3 Crucibles - Porcelain or disposable aluminum.

      4.5   Pasteur glass pipets -  1  ml,  disposable.

      4.6   Beakers - 400 ml.

      4.7   Vacuum or pressure filtration  apparatus.

            4.7.1 Buchner funnel.

            4.7.2 Filter paper - Whatman  No.  41 or equivalent.

      4.8   Kuderna-Danish (K-D) apparatus.

            4.8.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
      equivalent).   A ground  glass  stopper  is used to  prevent  evaporation  of
      extracts.

            4.8.2 Evaporation   flask   -   500   mL  (Kontes  K-570001-500   or
      equivalent).  Attach  to   concentrator  tube  with  springs,  clamps,   or
      equivalent.

            4.8.3 Snyder column  -  Three  ball  macro  (Kontes K-503000-0121  or
      equivalent).

            4.8.4 Snyder  column -  Two ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.8.5 Springs - 1/2 inch  (Kontes  K-662750 or equivalent).

      4.9   Boiling chips - Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).
                                   3550A  -  2                         Revision 1
                                                                September 1994

-------
      4.10  Water  bath  -   Heated,  with  concentric  ring  cover,  capable  of
temperature control (+ 5°C).  The batch should be used in a hood.

      4.11  Balance -  Top loading, capable of accurately weighing to the nearest
0.01 g.

      4.12  Vials  - 2  ml,  for  GC autosampler,  with Teflon lined screw caps or
crimp tops.

      4.13  Glass  scintillation  vials - 20 ml, with Teflon lined screw caps.

      4.14  Spatula - Stainless  steel or Teflon.

      4.15  Drying column  - 20 mm ID Pyrex  chromatographic  column  with Pyrex
glass wool at bottom.
      NOTE:
Fritted  glass  discs  are difficult  to decontaminate  after
highly  contaminated  extracts  have  been  passed  through.
Columns without frits may be  purchased.   Use  a small  pad of
Pyrex glass wool to retain the adsorbent.   Prewash the glass
wool pad with 50 ml of  acetone  followed by 50 ml of elution
solvent prior to packing the column with adsorbent.
      4.16  Syringe - 5 ml.
5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used in all  tests.  Unless
otherwise specified, it is intended that all inorganic reagents  shall conform to
the  specifications  of the  Committee on Analytical  Reagents  of  the  American
Chemical Society, where such specifications are  available.  Other grades may be
used, provided it is first ascertained  that the  reagent  is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.  All  references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.
      5.3   Sodium sulfate (granular,  anhydrous),  Na2S04.  Purify by heating at
400°C for 4 hours in a shallow tray, or by precleaning the sodium sulfate with
methylene chloride.  If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed,  demonstrating that there is no interference from
the sodium sulfate.

      5.4   Extraction solvents.

            5.4.1 Low concentration  soil/sediment  and aqueous  sludge  samples
      shall be  extracted  using a solvent system that gives optimum, reproducible
      recovery for  the  matrix/analyte combination  to be measured.   Suitable
      solvent choices are given in Table 1.
                                   3550A  -  3
                                                  Revision 1
                                              September 1994

-------
            5.4.2 Methylene   chloride:Acetone,   CH2C12:CH3COCH3   (1:1,   v:v).
      Pesticide quality or equivalent.

            5.4.3 Methylene chloride, CH2C12.  Pesticide quality or equivalent.

            5.4.4 Hexane, C6H14.  Pesticide quality or equivalent.

      5.5   Exchange solvents.

            5.5.1 Hexane, C6H14.  Pesticide quality or equivalent.

            5.5.2 2-Propanol, (CH3)2CHOH.   Pesticide  quality or equivalent.

            5.5.3 Cyclohexane, C6H12.  Pesticide quality or  equivalent.

            5.5.4 Acetonitrile, CH3CN.  Pesticide  quality or equivalent.

            5.5.5 Methanol, CH3OH.   Pesticide quality or equivalent.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material to this Chapter, Organic Analytes, Sec.
4.1.


7.0   PROCEDURE

      7.1   Sample handling

            7.1.1 Sediment/soil samples - Decant and  discard any water layer on
      a sediment sample.  Mix sample  thoroughly, especially  composited samples.
      Discard any foreign objects such as sticks,  leaves, and rocks.

                  7.1.1.2     Determine the dry  weight of  the sample  (Sec. 7.2)
            remaining after decanting.  Measurement of soil  pH may be required.

            7.1.2 Waste  samples  -  Samples consisting  of multiphases  must be
      prepared by the phase separation method in Chapter Two before extraction.
      This procedure is for solids only.

            7.1.3 Dry waste  samples  amenable to grinding  -  Grind or otherwise
      subdivide the waste so that it either  passes through  a  1 mm sieve or can
      be extruded  through  a 1 mm hole.   Introduce sufficient sample into the
      grinder to yield  at least 100 g after  grinding.

            7.1.4 Gummy,  fibrous  or  oily  materials  not  amenable  to grinding
      should  be  cut, shredded,  or  otherwise broken  up to  allow  mixing,  and
      maximum exposure of the sample  surfaces for  extraction.  The professional
      judgment  of the  analyst is  required   for  handling  of  these  difficult
      matrices.
                                   3550A - 4                        Revision  1
                                                                September  1994

-------
      7.2   Determination  of  percent dry  weight -  In certain  cases,  sample
results are desired based on a dry weight basis.  When such  data are desired, or
required, a portion of sample for this determination should be weighed  out at the
same time as the portion used for analytical determination.

      WARNING:    The drying  oven should  be  contained  in  a hood or vented.
                  Significant laboratory contamination may  result from drying a
                  heavily contaminated hazardous waste sample.

      However, samples known  or  suspected to contain  significant concentrations
of toxic, flammable, or explosive constituents should not be overdried because
of concerns for personal safety.  Laboratory discretion is advised.   It may be
prudent to delay overdrying of  the weighed-out portion until other analytical
results are available.

            7.2.1 Immediately after weighing the sample for  extraction, weigh 5-
      10 g of the sample into a tared crucible.  Determine the % dry weight of
      the sample by drying overnight at  105°C.   Allow to cool  in a desiccator
      before weighing:

            % dry weight = q of dry sample x 100
                              g of sample

      7.3   Extraction method  for samples expected to  contain  low concentrations
of organics and pesticides (< 20 mg/kg):

            7.3.1 The following step should be performed rapidly to avoid loss
      of the more  volatile  extractables.   Weigh approximately 30  g  of sample
      into a 400 ml beaker.  Record the weigh to the  nearest 0.1  g.  Nonporous
      or wet samples (gummy or clay type) that  do not  have a free-flowing sandy
      texture must  be  mixed with 60 g of  anhydrous sodium sulfate,  using  a
      spatula.  If required,  more sodium sulfate may  be added.   After addition
      of  sodium  sulfate, the  sample should  be free  flowing.   Add  1 ml  of
      surrogate standards to  all  samples,  spikes, standards, and  blanks  (see
      Method 3500 for details  on the  surrogate  standard solution and the matrix
      spike solution).   For  the sample in each  analytical  batch  selected for
      spiking, add  1.0 ml of the matrix  spiking  standard.  For base/neutral-acid
      analysis,  the amount added of the surrogates and matrix spiking  compounds
      should result in a final concentration of 100 ng/ni of each base/neutral
      analyte and 200 ng//LtL of each acid analyte in  the extract  to be analyzed
      (assuming a 1 juL injection).  If Method 3640,  Gel-Permeation Cleanup,  is
      to  be  used,  add   twice  the  volume  of  surrogates  and  matrix spiking
      compounds since half of the extract is  lost due to loading of the GPC
      column.   Immediately add 100 ml of 1:1  methylene chloride:acetone.

            7.3.2 Place the  bottom   surface  of the  tip  of  the  #207 3/4  in.
      disrupter horn about 1/2 in.  below the surface  of the solvent,  but above
      the sediment  layer.

            7.3.3 Extract ultrasonically for 3 minutes, with output control  knob
      set at  10  (full  power) and  with mode switch  on Pulse (pulsing  energy
                                  3550A - 5                         Revision 1
                                                                September 1994

-------
rather than  continuous  energy) and  percent-duty  cycle knob set  at 50%
(energy on 50% of time and off 50% of time).  Do not use microtip probe.

      7.3.4 Decant the extract  and filter it through Whatman No. 41 filter
paper (or equivalent) in a Buchner funnel that  is attached to a clean 500
ml filtration flask.   Alternatively,  decant  the extract into a centrifuge
bottle and centrifuge at low speed to remove particles.

      7.3.5 Repeat the extraction two or more  times with  two additional
100 ml portions of solvent.   Decant off the  solvent  after each ultrasonic
extraction.  On the  final  ultrasonic extraction,  pour the entire sample
into the  Buchner funnel  and rinse with extraction solvent.  Apply a vacuum
to  the  filtration flask,  and  collect  the  solvent extract.   Continue
filtration until all   visible solvent  is removed  from the  funnel,  but do
not attempt to completely dry the sample, as the continued application of
a  vacuum  may result   in  the  loss of some  analytes.   Alternatively,  if
centrifugation is used  in  Sec.  7.3.4, transfer  the entire sample to the
centrifuge bottle.  Centrifuge at low speed, and then decant the solvent
from the bottle.

      7.3.6 Assemble  a Kuderna-Danish (K-D)  concentrator (if necessary) by
attaching  a   10  ml  concentrator  tube  to   a  500  mL  evaporator  flask.
Transfer filtered extract to  a  500 ml evaporator flask  and proceed to the
next section.

      7.3.7 Add one to two clean boiling chips to the evaporation flask,
and attach a  three ball Snyder column.  Prewet the Snyder column by adding
about 1 ml methylene  chloride  to the top.   Place  the K-D  apparatus on a
hot water  bath (80-90  °C)  so  that  the  concentrator  tube  is  partially
immersed in  the  hot  water and  the  entire  lower rounded  surface  of the
flask is  bathed with hot  vapor.   Adjust  the vertical position  of the
apparatus  and  the   water  temperature,  as  required,  to  complete  the
concentration in 10-15 min.  At the proper rate of distillation the balls
of the column will  actively chatter,  but the chambers will not flood with
condensed  solvent.    When  the  apparent volume  of  liquid reaches  1 ml,
remove the K-D apparatus  and allow  it  to drain  and cool  for at least 10
min.

      7.3.8  If a solvent exchange is required (as indicated in Table 1),
momentarily  remove the  Snyder  column,  add  50  ml of the exchange solvent
and a new boiling chip,  and re-attach the Snyder column.  Concentrate the
extract as described  in  Sec.  7.3.10,  raising the temperature of the water
bath, if necessary,  to  maintain proper  distillation.   When the apparent
volume again reaches 1-2 ml,  remove  the  K-D apparatus and  allow it to
drain and cool for at least 10 minutes.

      7.3.9  Remove the  Snyder  column and rinse the  flask  and  its lower
joints into  the  concentrator tube with 1-2 ml  of  methylene chloride or
exchange solvent.   If sulfur crystals are  a  problem,  proceed  to Method
3660 for cleanup.  The  extract may  be further concentrated by using the
technique outlined in Sec. 7.3.10 or adjusted  to  10.0 ml with the solvent
last used.
                             3550A -  6                        Revision 1
                                                          September 1994

-------
            7.3.10      If further concentration is indicated in Table  1, either
      micro  Snyder column  technique  (Sec.  7.3.10.1)  or  nitrogen blow  down
      technique  (Sec.  7.3.10.2)  is  used  to adjust  the  extract to  the final
      volume required.

                  7.3.10.1    Micro Snyder Column Technique

                        7.3.10.1.1        Add a clean boiling  chip  and attach a
                  two ball micro Snyder column to the concentrator tube.  Prewet
                  the  column  by  adding  approximately  0.5  mL of  methylene
                  chloride  or exchange solvent  through  the  top.    Place  the
                  apparatus in the hot  water bath.   Adjust the  vertical position
                  and  the  water temperature,   as  required,  to complete  the
                  concentration  in  5-10  minutes.    At   the   proper  rate  of
                  distillation the balls  of  the column will actively chatter,
                  but the chambers will not flood.   When  the liquid reaches an
                  apparent volume of  approximately  0.5  ml, remove the apparatus
                  from the water bath  and allow to drain  and cool for at least
                  10 minutes.   Remove the micro  Snyder  column  and  rinse  its
                  lower joint with approximately 0.2 ml of appropriate solvent
                  and add to the concentrator tube.   Adjust  the  final volume to
                  the  volume required  for  cleanup  or for the determinative
                  method  (see Table 1).

                  7.3.10.2    Nitrogen Slowdown Technique

                        7.3.10.2.1        Place the concentrator tube  in a warm
                  water bath  (approximately  35 °C)  and  evaporate  the solvent
                  volume to the required level  using a gentle stream of clean,
                  dry nitrogen (filtered through a  column of activated carbon).

                        CAUTION:    Do not use plasticized  tubing  between  the
                                    carbon trap and the sample.

                        7.3.10.2.2        The internal  wall  of the tube must be
                  rinsed down several  times with the  appropriate solvent during
                  the operation.  During evaporation, the solvent level  in  the
                  tube must be positioned  to  prevent  water from condensing into
                  the sample (i.e., the solvent level should be  below the level
                  of the  water  bath).   Under normal  operating conditions,  the
                  extract should not  be allowed to become dry.

                        CAUTION:    When the  volume of  solvent  is reduced below
                                    1 ml,  semivolatile analytes may be lost.

      7.4   If analysis of the extract will not be performed immediately, stopper
the concentrator tube  and store refrigerated.   If the extract  will  be  stored
longer than 2 days, it should be transferred to a  vial  with a Teflon lined  cap
and labeled appropriately.
                                   3550A  -  7                         Revision 1
                                                                September 1994

-------
      7.5   Extraction method for samples expected to contain high concentrations
of organics (> 20 mg/kg):

            7.5.1 Transfer approximately 2 g (record weight to the nearest 0.1
      g) of sample to  a 20 ml vial.  Wipe the mouth  of the vial with a tissue to
      remove any sample material.  Record the exact  weight of  sample taken.  Cap
      the  vial  before  proceeding with  the next  sample   to  avoid  any  cross
      contamination.

            7.5.2 Add  2  g of  anhydrous  sodium sulfate to  sample  in  the  20 ml
      vial and mix well.

            7.5.3 Surrogate standards  are added to  all  samples,  spikes,  and
      blanks (see Method 3500 for details on the  surrogate standard solution and
      on the matrix spike solution).  Add 1.0 ml of surrogate spiking solution
      to sample mixture.  For the sample in each analytical batch selected for
      spiking, add 1.0 ml of the matrix spiking standard.  For base/neutral-acid
      analysis, the amount added of  the  surrogates  and matrix spiking compounds
      should result in a final  concentration of  100 ng//xL of each base/neutral
      analyte and 200 ng/^L of each acid analyte in the extract to be analyzed
      (assuming a 1 fj.1  injection).  If Method 3640, Gel-Permeation Cleanup, is
      to  be  used,  add  twice  the  volume  of  surrogates  and matrix  spiking
      compounds since half the extract is lost due to loading  of the GPC column.

            7.5.4 Immediately add whatever  volume  of solvent  is  necessary to
      bring  the  final  volume  to  10.0  ml  considering  the  added  volume  of
      surrogates and matrix  spikes.  Disrupt the  sample with the 1/8 in. tapered
      microtip ultrasonic probe for 2 minutes at output control  setting  5 and
      with  mode  switch on  pulse  and  percent duty  cycle  at  50%.   Extraction
      solvents are:

            1.    For  nonpolar compounds  (i.e.,  organochlorine  pesticides and
                  PCBs), use hexane or appropriate  solvent.

            2.    For extractable priority pollutants,  use  methylene chloride.

            7.5.5 Loosely pack disposable Pasteur  pipets with  2  to 3 cm Pyrex
      glass wool plugs.  Filter the extract through the glass wool and collect
      5.0  ml  in  a  concentrator  tube  if  further  concentration  is  required.
      Follow  Sec.  7.3.10 for  details  on concentration.  Normally,  the 5.0 mL
      extract is concentrated to approximately 1.0  ml or less.

            7.5.6 The extract  is ready for cleanup or analysis, depending on the
      extent of interfering co-extractives.


8.0   QUALITY CONTROL

      8.1   Any reagent  blanks or matrix  spike  samples  should be subjected to
exactly the same analytical  procedures as those  used on actual samples.
                                   3550A  -  8                         Revision 1
                                                                September 1994

-------
      8.2   Refer to  Chapter  One for specific  quality  control  procedures and
Method 3500 for extraction and sample preparation procedures.


9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative method for performance data.


10.0  REFERENCES

1.    U.S. EPA 40  CFR  Part 136, "Guidelines Establishing  Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act;  Final Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

2.    U.S.  EPA,  Inter!aboratory Comparison Study:   Methods  for  Volatile and
      Semi-Volatile Compounds,   Environmental  Monitoring  Systems  Laboratory,
      Office of Research  and Development, Las Vegas,  NV,  EPA 600/4-84-027, 1984.

3.    Christopher S. Hein,  Paul  J. Marsden, Arthur S. Shurtleff, "Evaluation of
      Methods 3540 (Soxhlet) and  3550 (Sonication) for Evaluation of Appendix IX
      Analytes form Solid Samples",  S-CUBED,  Report for EPA Contract 68-03-33-
      75, Work Assignment No.  03, Document No. SSS-R-88-9436, October 1988.
                                  3550A  - 9                         Revision 1
                                                                September 1994

-------
                   CO
                   5?
                   a
                   oo
   oo
   •>-
   oo
   O
   oo
                   a
                   oo
                   cc
                a
                oo
                   a:
                   o
                   oo
03
ov£>ooocMOc\jiooiorocor-.«d-ior--ojoc\i!—i CM

i—it—iLni—ic\ii—ii—iOcMCMcMoo^t-"3-i—(•—i«d-roooro

•«d- •—iLOOOOOOOLDOO'd-CriOOOi—i LT> CM «d- O IO O O

co «a-1— *i-i^.i»-cMO^»-co«3-CT>r--.cocMi—i^i-cooo
i^. co ro    LOIO        ai 10 10 in r-- r*- i—i       co •—i LO CM


**• CO ro r^. 10 o  •—iLr>Ln*ooOLncM«3-i— «3- o oo •—i o

o ro -si- •—i *»• i—ioo-tf-LnrocMCM>-HOO«a-O^i-ioeMr-~
         i—i        i—i  i—i                         i—i


Lf)'a-oocoor^e>joooc\j«d-cMOOrocT>iocricT»iOLno

•4- at «4- •—'poocovocnmooio^-ojwoior^i^oio^r
oo co i—vooocom^s-aiLniovxjcocoi— ^-mcovovooo


eyt«s-^-«d-CMmLncNjooLnc-~oOLr>i—iLn*j-r-»co^i-io

i—iMcvioococMioiooorO'—i^-i—icM^j-ior-^^j-iocooo

r^.^a-oocMOOiooOLO'—iCMioroi—-oo^i-mi—11--. en 10 »a-

«3 r-» CTI •—i o 10  r-~  o i—i^j-mootouoooLn'—i^j-iOLnoo
cocT>u3^oa>cMCMcMr^cacaoocr>ic>ooooi^«d-ic>r^
            i—(          i—i


m 00 «3- •—iCMLnoi~^rr>Lncr>r^ocMr~.caii>i^unooo

OCMinOOCOCMCvJ'—I O CSJ «* O •—I i—I CM i—l i—i >5l- LO CO **"


«a-rococvjo^ooor^roooiOroi~~LOOcoLnior-.ocri

io*rooi^-ooooLocMooroLncocov£>i—ia>Ln«3-cocMCT>
mi— ir>i—ir--.tD>—'•—icMioiOiOLOio^-i—ii—i 10 ro LT> ^


in *f in ao co •—iLOOOiootDioooooconoor-^ota-m

ioio^r^»-oooO'd-'d-i—lOO"—ir^.r-~oooorocoior--c>sjLn
                                                    c\j


c\ir-«cvjO'a-c\jrorooO'—icnoLD.—icooo«a-iocoior~-

^-vO'—icviiooocot—CID
o
>-
                   CO
                  .O

                    o
                   oo
                   OOO
                               I— ii— ii— iLnir>i— i^rn-i— iiD«t-cor-»r^i — ^
                               i— !•— icjiocn^s-oorocvjocvji— icxji — trj
                               i — It— I     O    Lf)    ID r— I IO O >— I
                             s-
                             cu
                             cu  o +->
                                c:  a>
          i_
          cu
                                            s_
                                            0)
                                            cu
                                      cu
'E
 CO

 00

 oo
             c
             cu
             o
             oo
                             c:  CL >>
                             Q) i—  X i
                            .c  >, o
                            i—  CO  CU
                             >> E  O
CU
c
cu
                                         C CU
                                         Q) C
                                         -c cu
                                         Q. N
                                            C
                                         i— cu
                          CO CO  O O
                         _C  I i— i—
                                  : .C
                c
                3
                O
                ex

                o
                o
 Q. O .C
 o  s- «_:>
 E  O  i
 O i— CVI
                                     CVI
                                         a.
                                         rt>
 c  o
 cu  s-
-C  O
 Q-r—
 O -C
 S-  U
 O -r-
r- Q
 O)

 CO


 CO
n
 o .

 o.

JC
 u I

Q .
                                            O              CO
                                            oo  CO  CO  CO     £=
                                          0) CO  C  C -O  CU  CD
                                                CU  CU ••-  C -r-
 S-
 CJ
 I
 o  o  o
                                                                  X  OJ T3
                                                                  O  N
                               o o

                               -t-J +->
                            CO
                             I
                                   oo
                                                     Q) ID
         •I—   I   I
         -Q CM ^J-
          Q. C +->
          CO  CO  3
             ^^ _.^^[

          S-  O  O
          O  S-  S-
v- -r- .r- ,—  O  O
 C  C  C .C i— i—
•r- T- •<-  O -C J=
Q Q Q  (T3  U  U
 i   i   i  -t_>  ro  ro
          Q. X  X
          CU  CU  CO
                                                                           cu
                                                                           c
                                                                           cu
+->     cu
 c     c:
 CU    -1-
 Q. CU T3
 O  C -r-
i—  n3  3
 O -C r—
 >»-!->  O  CO
 O  CO
 O  O
                                                                           S-  .
                                                                           o  o
                                                                          i— i—  O
   c
   co
O N
   c
   cu
                                                           CM CM re
                               o o +-> o
                               ro ro -r- S-
                               X X Z +->
                               CO CO  I  -r-
                               ic re uo z
                                                              cu
                                                              c
                                                              cu
                                                              N
                                                              c
                                                              a>
                                                             .a
                                                              o
                                                              S-
                                                              o
                                                                                       O «*•
                                                                                       c
                                                                                       CU OsJ
                                                                                                                                    o ^-<
                                                                                                                                 •r- Q)
                                                                                                                                 > JD
                                                                                                                                 cu E
                                                                                                                                 a: cu
                                                                                                  IT)
                                                                                                  «a-
                                                                                                  UD
                                                                                                  Di
                                                                                                  oo
                                                                                                cu
                                                                                                  T3
                                                                                                   cu
                                                                                                   oo
                                                                                               oo
                                                                                                   en
                                                                      CT>


                                                                      O  S-
                                                                      o  co
                                                                      CM -Q

                                                                      +->  3
                                                                                                         ro
                                                                                                         S-
                                                                                                                                 o
                                                                                                                                 LT)
                                                                                                                                 LO
                                                                                                                                 CO
                                                                                                         3
                                                                                                         CU
                                                                      -o  >,
                                                                      Q)  S_
                                                                      ^ -M
                                                                      •i—  00
                                                                      Q.-C-
                                                                      00  C7>
                                                                          ID
                                                                      00 C£
                                                                      a>
                                                                      +->  cu
                                                                                               ns  >
                                                                                               C  i-
                                                                                               m  cu
                                                                                                  00
                                                                cu
                                                                00

                                                               CO

                                                                II

                                                               CD
                                                                    cu
                                                                    c
                                                                    o
                                                                    -»J
                                                                    cu
                                                                    u
                                                                      ro
                                                         c  o
                                                         CU -r-
                                                         U  E
                                                         S-  CU
                                                         cu -c
                                                              o  cu
                                                             < -o
                                                                •r—
                                                              II  S-
                                                                 o
                                                                 o

                                                              cu  cu
                                                              a. c
                                                              >, cu
                                                                                                             o

                                                                                                             ns
                                                                                                            -C
                                                                                                            -I-J
                                                                                                             cu
                                                                                                            2:

                                                                                                     ,^-i co cu
                               cu

                              • r-	CU CU
                               s_
                               O  O) r— i—
                              i—  £:>>>,
                              JZ  O +J -(-»
                               U +-> 3 3
                                  CO -O -O
                               CU  O  i   i
                                                                                                            CU
-C  ns -E .c
+J  X +J +->
 CO  CO  CU  CU
                                                                                                    II  II  II  II

                                                                                                   CO O O LJJ

-------
O
O
UJ
O
   O


   
4-
O








CU
Ol-t-J
C C
05 CU
f~ >
O r—
X O
LU 00







CU
U) -f-*
c E
05 CU
r- >
U r—
X 0
LU 00






















CU
E
ir
• (^
>


-O
CU
i-
•I™" &M.
3 O
0-4-
CU
S-









-o
cu
i—
•r— 5^.
3 O

CU
s-







T3
CU
S-
•r- S-
3 O
CT4-
0)
S-

c
o
-M
(J
OS
S-
X
LU
CU
05
C
•f—
E
S-
0)
cu
Q

»•"»
_I_LIL-*
' 	
00

OO
^^
'oJ
E
03
E


O-
3
C
03
CU
P—
O








Q.
3
C
OJ
cu

o







00
•r-
00
>j
r_
OJ
c~
05




^p
O.




-0
0

+j
CU



_Q
o
o
r— 1
^
O 0
^ /™>
r—1 ^2





O 0
i—l CM














CU CD
C C
03 03
X X
cu cu
J= -c






o
c
03
O. CU
0 C
S- 03
QL X
i CU
CM J=
•O T3
CD CU
> >
03 CU
>
03 J=
X •»->
cu cu
j= e







P— •
o
cu c
C 03
03 J=
X -(->
cu cu

•o ~a
cu cu
> >
cu cu
0 e_>
cu cu
s- s-
oo oo
03 (D



i-i O
to r*«»
o o
00 CO







O 0
*—*. *—+.
^2 ^J





0 0
0 0
1 — 1 I— 1













CU CD
c. c
03 03
X X
cu at
J= -C









cu cu
E E
OS OS
X X
cu cu
J= J=
•a -o
cu CD
> >
CU O)
O U
cu cu
s- s-
00 oo
03 03



0 ^
OO 00
O O
00 00







0 0







0 0
CM CM









CU
E
OS
X
cu
CU -C
E 0
03 r—
X O
CU >>
J= O









CU
E
03 CU
X E
CU O
J= E
-o -o
CU CU
> >
cu a>
0 0
cu cu
S- S-
00 00
03 03



O O
C7> O
O r-l
00 00







0 0







O O
CM CM














CD CU
E E
03 OS
X X
Q) Q)
J= JC









CU CD
E E
OS 03
X X
CU CD
•^•^
-a -a
CD CU
> >
cu cu
o o
cu cu
S- S-
00 00
03 03



o o
•— < CM
i — t i— H
00 00







0 0







o
CM 1














cu
C"
03 i
X i
 >
O) CU
CJ O
cu cu
S- S-
00 01
03 03

u
(D
r-< O
CM CO
i— 1 CM
00 00







o








1
















1
1













cu
E
O
E
^
CU
>
cu
(J
cu
s-
00
03


u
O
f^
CM
00







0 O








1 1
















1 1
1 1






cu
r—
s-
+-> r—
•r- 0
C E
O 05
+•> -E
CU 4->
U CD
03 E
•a -o
CU CU
> >
cu cu
0 (J
CU CD
s- s-
oo oo
OJ O3



0 T->
i— i CM
oo PO
00 00



• — •
s-
•o

o







0
o





cu
-o
s-
o
J=
(J

cu
E
cu
r—
>^
t~
+->
cu
E
cu
T3
S-
o
J^
(J
cu
e~
CU
r—
>^
J=

CU
E
-o
cu
>
cu
(J
cu
s_
00
03



o
1 — 1
^^
00


                                                                                         c
                                                                                         o
                                                                                         OS
                                                                                         i.
                                                                                         +J
                                                                                         c
                                                                                         CU
                                                                                         u
                                                                                         c
                                                                                         o
                                                                                         u

                                                                                         Dl
                                                                                         C
                                                                                         •r-

                                                                                         O
                                                                                        T3
                                                                                         0)
                                                                                         S-
                                                                                         cu
                                                                                         Q-

                                                                                         cu
                                                                                        -Q
                                                                                         3
                                                                                         O
o
in

oo


 o
j=
4->
 CU
                                                                                         00
                                                                                         •M
                                                                                         03
                                                                                         S-
                                                                                         X
                                                                                         cu
                                                                                         03
                                                                                         S-
                                                                                         3
                                                                                         CU
 CD
 oo
 OJ
JO
 E
 03  —I
    E
•a
•r-  O
 O   •
 03  O
    r—I
 CU
+->  o
 03  +->
 S-
 OS  +->
 a. o
 cu  os
 00  S-

 E  X
•r-  CU
 03
+->  cu

 o  -»->

 O  <4-
           O  CU
           00  JO

          'oJ  o
              -M
          O

          StJ
          00
           O  X
          J=  CU
              03
              X
              CO
          o
          43
           O  «=
           03  -F~
           i.
          +->  oo
           X  4J
           CU  i—
              3
          i—  ^
           O  
                                                                                                                 cu

                                                                                                                 o
                                                                                                                S-
                                                                                                                CD
                                                                                                                cu
                                                                                                                a:
                                                                                                                0>
                                                                                                                oo
                                                                                                                00
                                                                                                                cu
                                                                                                                u
                                                                                                                CD
                                                                                                                c
                                                                                                                c
                                                                                                                3
                                                                                                                u
                                                                                                                03
                                                                                                                X
                                                                                                                cu
                                                                                                                                            cn
                                                                                                                                         c en
                                                                                                                                         O !-«

                                                                                                                                         V>  S-
                                                                                                                                         -r-  QJ
                                                                                                                                         > -Q
                                                                                                                                         CU  E
                                                                                                                                         C£  CU
                                                                                                                                            •(->
                                                                                                                                             a.
                                                                                                                                             cu
                                                                                                                                            oo
                                                                                                                cu
                                                o
                                                in
                                                in
                                                oo
          •r-  CU
           00  i.
           3  3
             -o
            ^ CU
          O  U
          ^f  O
          O  S-
          00  O-
                                                                                                   o o
                                                                                                  J= ••-
                                                                                                  +J •»->
                                                                                                   CU 03
                                                                                                  JD 03
                                                                                                  -o
                                                                                                   cu
                                                                                                   N
                                                                                                   ^S  (&

                                                                                                   S5i
                                                                                                   cu  Q.,
                                                                                                  -Q  o ,
                                                                                                   03 O3 j^

                                                                                                      oo -a
                                                                                                   oo c  cu
                                                                                                  r— -r-  M
                                                                                                   O OJ  >,
                                                                                                   C +•> r—
                                                                                                   O) C  03
                                                                                                  J= O  C
                                                                                                  a. o  oj
                                                                                                                    cu
                                                                                                                    3

                                                                                                                I?
                                                                                                                fty iji.
                                                                                                                cu ^1
                                                                                                                'U  Q,
                                                                                                                   ,__
                           00
                       en cu
                       s: s_

                       O 73
                       ts cu
                        O S-
                           a.
                        >>
                       4-> OL
                       •t- 3
                        U C
                       •r- OJ
                       <+- CU
                       •r* r— -
                        u o
                        cu
                        a. cu
                        cu

-------
                            METHOD  3550A
                      ULTRASONIC EXTRACTION
                                |     Start     j


                                         ~^
                                7.1 Prepare samples
                              ucing appropriate method
                                for the waete matrix
                                  7.2 Determine the
                                 percent dry weight
                                   of the eample
7.5.2 Add anhydroue
  sodium sulfate to
      sample
          7.5.2
        le organic
      concentration
     expected to be
       < 20 mg/kg?
7.3.1 Add eurrogate
  etandarde to all
  •ample*, epika*,
    and blanke
 7.5.3 Add eurrogate
   standards to all
   sample*, spikee,
     and blanke
                                      7.3.2 - 7.3.5
                                    Sonicate sample at
                                      least 3 times
        I
7.5.4
Adjust
volume; disrupt
sample with tapered
microtip
ultrasonic
probe


                                                                   7.3.7 Dry and
                                                                  collect extract in
                                                                  K-D concentrator
      7.5.5 Is
       further
    concentration
      required?
   7.5.5 Filter
through glass wool
  7.3.8 Concentrate
  extract and collect
 in K-D concentrator
                                                                     o
                             3550A  -  12
                                                Revision  1
                                           September  1994

-------
                      METHOD 3550A
                        continued
7.3.9 Add exchango
     solvent;
concentrate extract
Yes
 7.3.10 Use Method
  3660 for cleanup
                                      Yes
 7.3.9 Is
 a solvent
exchange
 required?
          7.3.10 Do
        sulfur crystals
            form?
                                                7.3.11 Further
                                              concentrate and/or
                                                adjust volume
                                             /  Cleanup or   \
                                             I    analyze    I
                        3550A -  13
                           Revision 1
                      September  1994

-------
                                 METHOD 3580A
                                WASTE DILUTION
1.0   SCOPE AND APPLICATION
      1.1   This method  describes  a solvent  dilution  of a  non-aqueous  waste
sample prior to  cleanup  and/or analysis.   It is designed for  wastes  that may
contain organic chemicals at a  concentration greater than 20,000 mg/kg and that
are soluble in the dilution solvent.
      1.2   It is recommended that an aliquot of the diluted sample be cleaned
up.  See this chapter, Organic Analytes, Section 4.2.2 (Cleanup).
2.0   SUMMARY OF METHOD
      2.1   One gram of sample is weighed  into a capped tube, and the sample is
diluted to 10.0 mL with an appropriate solvent.

3.0   INTERFERENCES
      3.1   Refer to Method 3500.
4.0   APPARATUS AND MATERIALS
      4.1   Glass scintillation vials:   At  least 20 mL, with Teflon or aluminum
foil lined screw-cap, or equivalent.
      4.2   Spatula:  Stainless steel or Teflon.
      4.3   Balance:  Capable of weighing 100 g to the nearest 0.01 g.
      4.4   Vials and caps:  2 mL for GC autosampler.
      4.5   Disposable pipets:  Pasteur.
      4.6   Test tube rack.
      4.7   Pyrex glass wool.
      4.8   Volumetric flasks, Class A:  10 mL (optional).
5.0   REAGENTS
      5.1   Sodium sulfate (granular, anhydrous),  Na2S04.  Purify by heating at
400°C for 4 hours in a shallow tray, or by  precleaning the sodium sulfate with
methylene chloride.   If the sodium sulfate is precleaned with methylene chloride,

                                  3580A -  1                         Revision 1
                                                                     July 1992

-------
a method blank must be analyzed, demonstrating that there is no interference from
the sodium sulfate.

      5.2   Methylene chloride,  CH2C12  - Pesticide quality or equivalent.

      5.3   Hexane, C6HU - Pesticide quality or equivalent.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Section 4.1.


7.0   PROCEDURE

      7.1   Samples  consisting of multiphases must  be prepared by  the  phase
separation method (Chapter Two)  before extraction.

      7.2   The sample dilution may be  performed in  a  10 mL volumetric flask.
If disposable  glassware is  preferred, the  20  mL scintillation  vial may  be
calibrated for use.  Pipet  10.0 mL of extraction solvent  into the scintillation
vial  and mark the bottom of the meniscus.   Discard  this solvent.

      7.3   Transfer approximately 1 g of each phase  of the  sample to separate
20 mL vials or  10  mL volumetric  flasks (record weight  to the  nearest 0.1 g).
Wipe the mouth  of the vial with a tissue to remove any sample material.  Cap the
vial  before proceeding with the next sample to avoid  any cross-contamination.

      7.4   Add 2.0 mL surrogate spiking solution to all samples and blanks.  For
the sample  in  each analytical batch selected  for  spiking,  add 2.0  mL of the
matrix spiking  standard.  For base/neutral-acid analysis, the amount added of the
surrogates and matrix spiking compounds should result in a final concentration
of 200 ng//iL of each base/neutral  analyte  and 400 ng/^L of each acid analyte in
the extract to be  analyzed  (assuming a  1  /uL  injection).  If Method 3640,  Gel-
permeation cleanup, is to be used, add twice the volume  of surrogates and matrix
spiking  compounds  since half the extract is lost due to loading  of the GPC
column.  See Method 3500 and the determinative method to  be  used for details on
the surrogate standard and matrix spiking solutions.

      7.5   Immediately dilute  to 10  mL  with the appropriate solvent.   For
compounds to be analyzed by GC/ECD, e.g., organochlorine pesticides  and PCBs, the
dilution  solvent  should be  hexane.    For  base/neutral  and  acid  semivolatile
priority pollutants, use methylene chloride.   If the  dilution is to be cleaned
up by gel permeation chromatography (Method 3640),  use methylene  chloride as the
dilution solvent for all compounds.

      7.6   Add 2.0 g of anhydrous sodium sulfate to  the sample.

      7.7   Cap and  shake the sample for 2 min.
                                   3580A -  2                         Revision 1
                                                                     July 1992

-------
      7.8   Loosely pack disposable Pasteur pipets with 2-3 cm glass wool plugs.
Filter the extract through the glass wool and collect 5 mL of the extract in a
tube or vial.

      7.9   The extract is ready for cleanup or analysis, depending on the extent
of interfering co-extractives.


8.0   QUALITY CONTROL

      8.1   Any reagent blanks and matrix spike samples should be subjected to
exactly the same analytical procedures as those used on actual samples.

      8.2   Refer to  Chapter  One for specific quality  control  procedures and
Method 3500 for extraction and sample preparation procedures.


9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods for performance data.


10.0  REFERENCES

      10.1  None applicable.
                                  3580A - 3                         Revision 1
                                                                     July 1992

-------
                       METHOD  3580A
                      WASTE  DILUTION
7  3  Transfer 1 g o
  each phase to
 separate vials or
      flasks
 7 4  Add sur ragate
spiking solution to
  all  samples and
      blanks
  7 4  Add ma t rix
spiking  standard to
sample selected for
      spiking
  7 5 Dilute with
appropriate solvent
                           7  1 Use phase
                         separation method
                            (Chapter 2)
 7  6 Add anhydrous
 ammonium sulfa Le
 7  7 Cap and shake
7 8  Filler through
    glass wool
Cleanup or analyze
                            3580A -  4
                      Revision  1
                       July  1992

-------
                                 METHOD 5030A

                                PURGE-AND-TRAP
1.0   SCOPE AND APPLICATION
      1.1   This method  describes sample  preparation  and extraction  for  the
analysis  of  volatile  organics  by   a  purge-and-trap  procedure.    The  gas
chromatographic determinative steps are  found  in Methods 8010, 8015, 8020, 8021
and 8030.   Although applicable to  Methods 8240 and 8260,  the  purge-and-trap
procedure is already incorporated into Methods 8240 and 8260.

      1.2   Method 5030 can be used for most volatile organic compounds that have
boiling  points  below  200°C  and  are  insoluble or  slightly soluble  in  water.
Volatile water-soluble compounds can be included in this analytical technique;
however, quantitation limits  (by GC or GC/MS) are approximately ten times higher
because of poor purging efficiency. The method is also limited to compounds that
elute as  sharp peaks from a  GC column packed with  graphitized  carbon  lightly
coated with  a carbowax  or  a coated capillary column.   Such compounds include low
molecular  weight  halogenated  hydrocarbons,   aromatics,   ketones,  nitriles,
acetates, acrylates, ethers,  and sulfides.

      1.3   Water samples  can be analyzed directly for volatile organic compounds
by purge-and-trap extraction  and gas chromatography.  Higher concentrations of
these analytes in water can be determined by direct  injection of the sample into
the chromatographic  system.

      1.4   This method also describes the preparation of water-miscible liquids,
non-water-miscible liquids, solids, wastes,  and soils/sediments for analysis by
the purge-and-trap procedure.


2.0   SUMMARY OF METHOD

      2.1   The  purge-and-trap  process:   An inert  gas is  bubbled  through  the
solution  at  ambient  temperature,  and the  volatile  components  are  efficiently
transferred  from the aqueous phase  to  the vapor phase.   The vapor  is swept
through  a sorbent  column where the  volatile  components are adsorbed.   After
purging is completed, the sorbent column  is  heated and backflushed with inert gas
to desorb the components onto a gas chromatographic column.

      2.2   If  the  sample  introduction   technique in  Section  2.1   is  not
applicable,  a portion  of  the sample is dispersed in methanol  to  dissolve  the
volatile organic  constituents.  A portion of the methanolic  solution is combined
with water  in  a  specially designed  purging chamber.   It  is then  analyzed by
purge-and-trap GC following the normal water method.


3.0   INTERFERENCES

      3.1   Impurities in the purge gas, and from organic compounds out-gassing
from the plumbing ahead of the trap,  account for the majority of contamination
problems.    The  analytical  system  must  be demonstrated  to  be  free  from

                                   5030A -  1                         Revision 1
                                                                     July 1992

-------
contamination under the conditions of the analysis by running laboratory reagent
blanks.  The use of non-TFE  plastic  coating,  non-TFE  thread  sealants,  or flow
controllers with rubber components in the purging device should be avoided.

      3.2   Samples can  be  contaminated by  diffusion  of volatile  organics
(particularly methylene chloride and fluorocarbons) through the septum seal of
the  sample  vial during  shipment and  storage.   A trip  blank prepared  from
organic-free reagent water and carried through sampling and handling protocols
serves as a check on such contamination.

      3.3   Contamination by  carryover can occur whenever high-concentration and
low-concentration  samples  are analyzed  sequentially.   Whenever  an unusually
concentrated  sample  is  analyzed,  it  should  be  followed by  an  analysis  of
organic-free reagent water to check for cross-contamination. The trap and other
parts of the system are subject to contamination.  Therefore,  frequent bake-out
and purging of the entire system may be required.

      3.4   The  laboratory  where  volatile  analysis  is  performed  should  be
completely free of solvents.


4.0   APPARATUS AND MATERIALS

      4.1   Microsyringes - 10 jiL, 25 p.1,  100 /xL,  250 /LtL, 500 /iL, and 1,000 /uL.
These syringes should  be  equipped with  a 20 gauge  (0.006 in ID) needle having a
length sufficient to extend  from the sample inlet to  within  1  cm  of the glass
frit in the purging device.  The needle length will depend upon the dimensions
of the purging device employed.

      4.2   Syringe valve - Two-way,  with Luer  ends (three  each), if applicable
to the purging device.

      4.3   Syringe -  5 ml, gas-tight with shutoff valve.

      4.4   Analytical balance - 0.0001 g.

      4.5   Top-loading balance - 0.1 g.

      4.6   Glass scintillation  vials -  20 ml,  with screw-caps and Teflon liners
or glass culture tubes with  screw-caps and Teflon liners.

      4.7   Volumetric flasks, Class A - 10 mL  and 100  ml_,  with  ground-glass
stoppers.

      4.8   Vials - 2 ml, for GC autosampler.

      4.9   Spatula - Stainless steel.

      4.10  Disposable pipets - Pasteur.

      4.11  Purge-and-trap device:  The purge-and-trap device consists of three
separate pieces  of equipment: the sample purger,  the  trap,  and the desorber.
Several complete devices are  commercially available.


                                   5030A - 2                         Revision  1
                                                                     July  1992

-------
      4.11.1      The recommended purging  chamber  is designed to accept 5
ml samples with a water column  at least 3  cm deep.  The gaseous headspace
between the water column  and  the trap must have  a total  volume of less
than 15 ml.  The purge  gas  must  pass  through  the  water column as finely
divided bubbles  with  a diameter of less  than  3  mm at the  origin.   The
purge gas must be introduced no more than  5 mm  from the base of the water
column.  The sample purger,  illustrated in  Figure 1,  meets these design
criteria.  Alternate sample purge devices may be used, provided equivalent
performance is demonstrated.

      4.11.2      The trap must be at  least  25  cm  long and have an inside
diameter of at  least  0.105  in.  Starting from the inlet,  the trap must
contain the following  amounts of adsorbents:  1/3 of 2,6-diphenylene oxide
polymer,  1/3   of silica  gel,  and 1/3   of  coconut  charcoal.     It  is
recommended that 1.0 cm of methyl silicone-coated packing be inserted at
the inlet to extend the life of the trap (see  Figures 2 and 3).  If it is
not necessary  to analyze for  dichlorodifluoromethane or  other fluoro-
carbons of  similar  volatility, the charcoal  can  be eliminated  and  the
polymer increased to  fill 2/3 of  the trap.    If  only  compounds boiling
above 35°C are to be  analyzed, both  the  silica gel and  charcoal  can be
eliminated and  the  polymer increased to  fill  the entire  trap.   Before
initial  use,   the  trap  should  be conditioned  overnight  at  180°C  by
backflushing with an inert gas  flow of at  least 20 mL/min.  Vent the trap
effluent to the hood,  not to the analytical  column.  Prior to daily use,
the trap should be conditioned  for  10 min at 180°C  with backflushing.  The
trap may  be vented  to the analytical  column  during  daily conditioning;
however, the column must be run through the temperature program prior to
analysis of samples.

      4.11.3      The desorber should be  capable  of  rapidly heating  the
trap to 180°C  for desorption.   The  polymer section of the trap should not
be heated higher than  180°C, and  the remaining  sections should not exceed
220°C  during bake-out  mode.  The  desorber  design illustrated in Figures 2
and 3 meet these criteria.

      4.11.4      The purge-and-trap device may be assembled as a separate
unit or  may  be  coupled  to  a  gas  chromatograph,  as  shown  in  Figures 4
and 5.

      4.11.5      Trap Packing Materials

            4.11.5.1    2,6-Diphenylene  oxide  polymer  -  60/80  mesh,
      chromatographic grade (Tenax GC  or equivalent).

            4.11.5.2    Methyl   silicone   packing   -   OV-1    (3%)   on
      Chromosorb-W, 60/80 mesh or equivalent.

            4.11.5.3    Silica gel  -  35/60 mesh,  Davison, grade  15 or
      equivalent.

            4.11.5.4    Coconut charcoal  - Prepare from Barnebey Cheney,
      CA-580-26  lot #M-2649, or  equivalent, by crushing  through 26 mesh
      screen.
                             5030A -  3                         Revision 1
                                                               July 1992

-------
      4.12  Heater  or  heated oil  bath  -  capable  of maintaining  the purging
chamber to within 1°C,  over a temperature range from ambient to 100°C.


5.0   REAGENTS

      5.1   Organic-free reagent water - All  references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.2   Methanol, CH3OH - Pesticide  quality or equivalent.  Store away from
other solvents.

      5.3   Reagent Tetraglyme - Reagent tetraglyme  is defined  as tetraglyme in
which interference is not observed at the method detection limit of the compounds
of interest.

            5.3.1 Tetraglyme (tetraethylene glycol dimethyl  ether, Aldrich #17,
      240-5 or equivalent), CgH18Or.   Purify by  treatment at  reduced pressure in
      a rotary  evaporator.   The tetraglyme should  have a  peroxide content of
      less  than  5 ppm  as  indicated by  EM Quant  Test  Strips  (available from
      Scientific Products Co., Catalog No. P1126-8 or equivalent).

            CAUTION:    Glycol  ethers  are suspected carcinogens.  All  solvent
                        handling should  be done in a hood  while  using proper
                        protective equipment to minimize exposure to liquid and
                        vapor.

            Peroxides may be removed by  passing the  tetraglyme  through a column
      of activated  alumina.  The tetraglyme is placed in a  round  bottom flask
      equipped with a standard taper joint, and the flask is  affixed to a rotary
      evaporator. The flask is immersed  in  a water bath at 90-100°C  and a vacuum
      is maintained  at  <  10  mm Hg  for at  least  two hours using  a  two stage
      mechanical pump.  The  vacuum  system is  equipped  with  an all  glass trap,
      which is  maintained  in a  dry  ice/methanol bath.  Cool  the  tetraglyme to
      ambient temperature and add 100 mg/L  of 2,6-di-tert-butyl-4-methyl-phenol
      to prevent peroxide  formation. Store the tetraglyme  in a tightly sealed
      screw cap bottle  in an area that  is not contaminated by  solvent vapors.

            5.3.2 In order to demonstrate that all  interfering volatiles have
      been removed from  the tetraglyme, an organic-free reagent water/tetraglyme
      blank must be  analyzed.

      5.4   Polyethylene glycol, H(OCH2CH2)nOH.  Free of  interferences  at the
detection limit of the  analytes.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Refer to the introductory material  to this chapter,  Organic Analytes,
Section 4.1. Samples should be stored in capped bottles,  with minimum headspace,
at 4°C or less.
                                   5030A -  4                        Revision 1
                                                                     July 1992

-------
7.0   PROCEDURE

      7.1   Initial calibration:  Prior to using this introduction technique for
any GC method, the system must be calibrated.  General calibration procedures are
discussed in Method 8000, while the  specific determinative  methods and Method
3500 give details on preparation of standards.

            7.1.1 Assemble a purge-and-trap device that meets the specification
      in Section 4.10.   Condition  the trap overnight at  180°C in the purge mode
      with an inert gas  flow of at least 20 mL/min.  Prior to use, condition the
      trap daily for 10 min while backflushing at 180°C with the  column  at 220°C.

            7.1.2 Connect the purge-and-trap device to a gas chromatograph.

            7.1.3 Prepare   the   final   solutions   containing   the   required
      concentrations of  calibration  standards,  including  surrogate standards,
      directly in the  purging device.  Add 5.0 ml of organic-free reagent water
      to the purging device.   The organic-free reagent water  is  added to the
      purging device using  a 5  ml glass syringe fitted with a 15 cm 20-gauge
      needle.   The  needle  is  inserted  through the  sample  inlet   shown  in
      Figure 1.  The  internal  diameter  of the  14-gauge needle that  forms the
      sample inlet will  permit insertion of the  20-gauge needle.  Next, using a
      10 jtL or  25  /it  micro-syringe  equipped  with a long needle (Section 4.1),
      take a volume of  the  secondary dilution solution  containing appropriate
      concentrations  of  the  calibration  standards.    Add   the  aliquot  of
      calibration  solution  directly  to  the organic-free reagent  water in the
      purging device by  inserting  the needle through the sample  inlet.   When
      discharging the  contents of the micro-syringe, be sure  that the end of the
      syringe needle  is  well beneath the  surface of  the  organic-free reagent
      water.  Similarly,  add 10 /xL of the internal standard solution.   Close the
      2-way syringe valve at the sample inlet.

            7.1.4 Carry  out the purge-and-trap  analysis  procedure  using the
      specific conditions given in Table 1.

            7.1.5 Calculate  response  factors or calibration factors  for each
      analyte of interest using the procedure described in Method 8000.

            7.1.6 The average RF must  be calculated for each  compound.  A system
      performance check  should  be made  before this  calibration curve is used.
      If the purge-and-trap procedure is  used with  Method  8010, the  following
      five  compounds   are  checked  for  a  minimum average  response factor:
      chloromethane; 1,1-dichloroethane; bromoform; 1,1,2,2-tetrachloroethane;
      and chlorobenzene.  The minimum acceptable average RF for these  compounds
      should be 0.300  (0.250 for bromoform).  These compounds typically have RFs
      of 0.4-0.6,  and  are  used to  check compound stability and  to  check for
      degradation caused  by contaminated  lines  or  active  sites in the system.
      Examples of these occurrences are:

                  7.1.6.1     Chloromethane:  This compound is the most likely
            compound to be  lost if the purge flow is too fast.

                  7.1.6.2     Bromoform:  This compound  is one  of the  compounds
            most likely to be purged very poorly if the purge flow  is  too slow.

                                   5030A - 5                         Revision 1
                                                                      July 1992

-------
            Cold spots and/or active sites in the transfer lines may adversely
            affect response.

                  7.1.6.3     Tetrachloroethane and 1,1-dichloroethane:  These
            compounds are degraded  by contaminated transfer lines in purge-and-
            trap systems and/or active sites in trapping materials.

      7.2   On-going calibration:  Refer to Method 8000 for details on continuing
calibration.

      7.3   Sample preparation

            7.3.1 Water samples

                  7.3.1.1     Screening of the sample  prior  to purge-and-trap
            analysis  will  provide guidance  on  whether sample  dilution  is
            necessary  and will  prevent contamination  of the  purge-and-trap
            system.  Two screening  techniques  that can  be utilized are: the use
            of an automated  headspace sampler (modified  Method 3810), interfaced
            to  a gas  chromatograph (GC),  equipped with  a  photo  ionization
            detector (PID),  in series with an electrolytic conductivity detector
            (HECD); and extraction of the sample with hexadecane (Method 3820)
            and analysis of the extract on a GC with a FID and/or an ECD.

                  7.3.1.2     All samples and standard solutions must be allowed
            to warm to ambient temperature before analysis.

                  7.3.1.3     Assemble the purge-and-trap device. The operating
            conditions  for  the GC  are  given in  Section  7.0 of  the  specific
            determinative method to be employed.

                  7.3.1.4     Daily GC calibration criteria must be met (Method
            8000) before analyzing samples.

                  7.3.1.5     Adjust  the purge  gas  flow rate  (nitrogen  or
            helium)  to that shown  in  Table 1,  on  the purge-and-trap device.
            Optimize   the   flow   rate  to  provide  the  best  response  for
            chloromethane  and bromoform,   if  these compounds are  analytes.
            Excessive   flow rate   reduces   chloromethane  response,   whereas
            insufficient flow reduces bromoform response.

                  7.3.1.6     Remove the plunger from a 5  ml syringe and attach
            a closed syringe valve.  Open the sample or standard bottle, which
            has been allowed to come to ambient temperature, and carefully pour
            the  sample into the  syringe barrel to  just short  of overflowing.
            Replace  the syringe plunger  and  compress  the  sample.   Open the
            syringe valve and vent any residual air while adjusting the sample
            volume to  5.0 ml.  This process of taking  an aliquot destroys the
            validity  of the liquid sample  for  future  analysis; therefore,  if
            there is only one VOA vial, the analyst should fill  a second syringe
            at this time  to protect against possible loss of sample integrity.
            This  second sample  is maintained only  until  such time  when the
            analyst  has determined  that the first sample  has  been  analyzed
            properly.  Filling one  20 ml syringe would allow the use of only one

                                   5030A - 6                        Revision  1
                                                                     July 1992

-------
syringe.  If a second analysis  is needed from a syringe, it must be
analyzed within  24 hr.   Care  must be taken  to prevent  air from
leaking into the syringe.

      7.3.1.7     The  following  procedure   is   appropriate  for
diluting purgeable  samples.   All  steps must  be  performed without
delays until the diluted sample is in a gas-tight syringe.

            7.3.1.7.1   Dilutions may be made in volumetric flasks
      (10 ml  to  100 ml).   Select  the volumetric  flask  that will
      allow for the necessary dilution.  Intermediate dilutions may
      be necessary for extremely large dilutions.

            7.3.1.7.2   Calculate   the   approximate   volume   of
      organic-free reagent water to be added to the volumetric flask
      selected and add slightly less than  this quantity of organic-
      free reagent water to the flask.

            7.3.1.7.3   Inject the proper aliquot  of  samples from
      the  syringe prepared  in Section   7.3.1.5  into the  flask.
      Aliquots of less than 1  ml  are  not  recommended.   Dilute the
      sample to the mark with organic-free reagent water.  Cap the
      flask,  invert,  and  shake  three  times.    Repeat  the  above
      procedure for additional  dilutions.

            7.3.1.7.4   Fill a  5 ml  syringe with the diluted sample
      as in Section 7.3.1.5.

      7.3.1.8     Add 10.0 /uL of surrogate spiking solution (found
in each determinative method, Section 5.0)  and, if applicable, 10 /xL
of internal  standard spiking solution  through  the valve bore of the
syringe; then close the valve.  The surrogate and internal standards
may be mixed and  added as a single spiking  solution.  Matrix spiking
solutions, if  indicated, should be added  (10  /zL)  to the sample at
this time.

      7.3.1.9     Attach the syringe-syringe valve assembly to the
syringe valve  on the purging device.   Open the  syringe valves and
inject the sample into the purging chamber.

      7.3.1.10    Close both valves  and  purge the  sample for the
time and at the temperature specified in Table 1.

      7.3.1.11    At the conclusion of the purge time, attach the
trap to the  chromatograph, adjust the device to the desorb mode, and
begin  the gas  chromatographic temperature  program  and GC  data
acquisition.  Concurrently, introduce the trapped materials to the
gas chromatographic  column by rapidly  heating  the trap to  180°C
while backflushing the trap with  inert gas between 20 and 60 mL/min
for the time specified in Table 1.

      7.3.1.12    While the  trap  is  being desorbed into  the gas
chromatograph, empty the purging chamber.   Wash  the chamber with a
minimum  of  two  5  ml  flushes  of organic-free  reagent  water (or

                       5030A -  7                         Revision 1
                                                         July 1992

-------
      methanol  followed  by organic-free reagent water) to avoid carryover
      of pollutant compounds into subsequent analyses.

            7.3.1.13    After desorbing  the sample,  recondition the trap
      by returning the purge-and-trap device to the  purge mode.  Wait 15
      sec;  then close  the syringe valve on the purging device to begin gas
      flow through the trap.  The trap temperature should be maintained at
      180°C for Methods 8010,  8020,  8021,  8240  and 8260 and  210°C  for
      Methods  8015  and 8030.    Trap temperatures  up  to  220°C may  be
      employed.  However, the higher temperatures will shorten the useful
      life of  the  trap.   After  approximately  7  min,  turn off  the trap
      heater and  open  the  syringe valve  to stop the  gas flow through the
      trap.  When cool,  the trap is ready for the next  sample.

            7.3.1.14    If the initial analysis of a sample or a dilution
      of the  sample has a  concentration of analytes  that exceeds  the
      initial calibration range,  the sample must be reanalyzed at a higher
      dilution.  When a sample is analyzed that has saturated response
      from a compound, this  analysis must be followed by a blank organic-
      free reagent water analysis.  If the blank analysis is not free of
      interferences, the system must be  decontaminated.  Sample analysis
      may  not  resume  until  a  blank  can be  analyzed  that  is free  of
      interferences.

            7.3.1.15    All  dilutions should keep  the response  of  the
      major constituents (previously saturated peaks) in the upper half of
      the  linear  range  of the  curve.   Proceed  to  Method 8000  and  the
      specific determinative method  for details on  calculating analyte
      response.

      7.3.2 Water-miscible liquids:

            7.3.2.1     Water-miscible  liquids  are  analyzed  as  water
      samples after first diluting them at least 50-fold with organic-free
      reagent water.

            7.3.2.2     Initial and serial dilutions can  be  prepared by
      pipetting 2  ml  of the sample  into  a 100 mL volumetric  flask and
      diluting to volume  with  organic-free  reagent  water.    Transfer
      immediately to a 5 ml gas-tight syringe.

            7.3.2.3     Alternatively, prepare dilutions directly in a 5
      ml syringe  filled with organic-free reagent water  by adding at least
      20 /iL, but  not  more  than 100 /iL of liquid sample.   The sample is
      ready  for  addition  of surrogate and,  if  applicable,  internal  and
      matrix spiking standards.

      7.3.3 Sediment/soil  and  waste  samples:  It is highly recommended
that all samples of this type be screened prior to the  purge-and-trap GC
analysis.   These  samples  may  contain  percent quantities  of  purgeable
organics that  will contaminate  the  purge-and-trap  system,  and require
extensive  cleanup and  instrument downtime.   See  Section 7.3.1.1  for
recommended  screening techniques.   Use  the  screening  data  to determine
whether to use the low-concentration method (0.005-1 mg/kg) or the high-

                             5030A -  8                        Revision 1
                                                               July 1992

-------
concentration method (>1 mg/kg).

            7.3.3.1     Low-concentration method:  This  is  designed for
      samples containing individual purgeable compounds of <1 mg/kg.  It
      is limited to sediment/soil samples and waste that is of a similar
      consistency (granular and porous).  The low-concentration method is
      based  on  purging   a   heated   sediment/soil   sample   mixed  with
      organic-free  reagent  water  containing  the  surrogate  and,  if
      applicable,  internal  and matrix  spiking  standards.    Analyze all
      reagent  blanks  and  standards  under  the   same  conditions  as the
      samples.

                  7.3.3.1.1   Use  a   5  g   sample   if  the   expected
            concentration  is  <0.1 mg/kg or  a  1  g sample  for  expected
            concentrations between 0.1 and 1 mg/kg.

                  7.3.3.1.2   The  GC   system  should  be  set up  as  in
            Section 7.0 of the specific determinative method.  This should
            be done prior to the preparation of the sample to avoid loss
            of  volatiles  from  standards  and  samples.   A  heated purge
            calibration  curve  must   be  prepared   and   used   for  the
            quantitation   of   all  samples   analyzed  with  the  low-
            concentration  method.     Follow  the   initial   and   daily
            calibration instructions,   except  for the addition  of a 40°C
            purge temperature for Methods 8010,  8020, and 8021.

                  7.3.3.1.3   Remove the plunger from a 5  mL Luerlock type
            syringe  equipped  with  a  syringe  valve  and  fill   until
            overflowing with  organic-free  reagent water.    Replace the
            plunger and compress  the  reagent  water to vent  trapped air.
            Adjust the  volume to 5.0  ml.   Add  10 p.1 each  of surrogate
            spiking solution and internal standard solution to the syringe
            through the valve.  (Surrogate spiking solution and internal
            standard solution may be  mixed  together.)   Matrix spiking
            solutions,  if indicated, should be added (10 juL)  to the sample
            at this time.

                  7.3.3.1.4   The sample (for  volatile organics) consists
            of  the entire  contents  of  the  sample  container.    Do not
            discard any  supernatant  liquids.    Mix the  contents  of the
            sample container  with a  narrow metal  spatula.   Weigh the
            amount determined in Section 7.3.3.1.1  into a  tared purge
            device.  Note and record the actual  weight to the nearest 0.1
            9-

                  7.3.3.1.5   Determination of  sample %  dry weight -  In
            certain cases, sample results are  desired based on dry weight
            basis.  When  such data is desired,   a portion of sample for
            this determination should  be weighed out at the same time as
            the portion used for analytical  determination.

                  WARNING:    The drying oven should be contained  in a
                              hood or vented.    Significant  laboratory
                              contamination may result  from a  heavily

                             5030A - 9                         Revision 1
                                                               July 1992

-------
                        contaminated hazardous waste sample.

                  7.3.3.1.5.1 Immediately after weighing the sample
            for extraction,  weigh  5-10 g of the sample into a tared
            crucible.   Determine the % dry weight of the sample by
            drying  overnight  at   105°C.    Allow  to  cool  in  a
            desiccator before weighing:

                  % dry weight = q of dry sample  x 100
                                    g of sample

            7.3.3.1.6    Add the spiked organic-free reagent water to
      the purge device, which contains the weighed amount of sample,
      and connect the  device to the purge-and-trap system.

            NOTE: Prior to  the  attachment  of the purge  device,
                  Sections   7.3.3.1.4   and   7.3.3.1.6   must   be
                  performed  rapidly and  without  interruption  to
                  avoid loss  of volatile organics.   These  steps
                  must be  performed in a laboratory free of solvent
                  fumes.

            7.3.3.1.7    Heat the sample to 40°C ±  1°C (Methods 8010,
      8020 and 8021) or to  85°C ±  26C (Methods 8015  and 8030)  and
      purge the sample for the time shown in Table 1.

            7.3.3.1.8    Proceed with  the  analysis as outlined in
      Sections 7.3.1.11-7.3.1.15.   Use 5 ml of the same organic-free
      reagent water as  in  the reagent blank.  If  saturated  peaks
      occurred or would occur if  a 1 g sample were  analyzed,  the
      high-concentration method must be followed.

            7.3.3.1.9    For    matrix    spike    analysis    of
      low-concentration sediment/soils,  add 10  \i\. of  the  matrix
      spike solution to 5 ml of organic-free reagent water (Section
      7.3.3.1.3  ).  The concentration for  a  5 g  sample would be
      equivalent to 50 jig/kg of each matrix spike standard.

      7.3.3.2     High-concentration method:  The method  is based on
extracting the  sediment/soil  with  methanol.   A waste  sample is
either  extracted  or   diluted, depending  on  its  solubility  in
methanol.    Wastes  (i.e.  petroleum  and coke  wastes)  that  are
insoluble  in  methanol  are  diluted  with  reagent tetraglyme  or
polyethylene glycol (PEG).   An aliquot  of the  extract is added to
organic-free reagent water  containing  surrogate and, if applicable,
internal  and matrix  spiking  standards.   This  is purged at  the
temperatures indicated  in  Table 1.  All  samples with an expected
concentration of >1.0 mg/kg should be analyzed by this method.

            7.3.3.2.1    The sample (for volatile organics) consists
      of  the entire  contents of  the sample  container.   Do  not
      discard  any  supernatant liquids.    Mix  the  contents  of  the
      sample   container  with  a  narrow  metal   spatula.     For
      sediment/soil and waste that are insoluble  in methanol, weigh

                      5030A  -  10                        Revision 1
                                                         July 1992

-------
4 g (wet weight)  of  sample  Into a tared 20 ml  vial.   Use a
top-loading balance.  Note and record the actual  weight to 0.1
gram and determine the  percent dry weight of the sample using
the procedure in Section 7.3.3.1.5.  For waste that is soluble
in methanol, tetraglyme,  or PEG, weigh 1 g (wet weight) into
a  tared scintillation  vial  or culture  tube  or  a   10  ml
volumetric flask.   (If a vial  or  tube is used,  it must  be
calibrated prior to use.   Pipet 10.0 ml of methanol  into the
vial  and  mark the  bottom  of  the  meniscus.    Discard this
solvent.)

      7.3.3.2.2   For sediment/soil or  solid waste,  quickly
add 9.0 ml  of appropriate  solvent; then  add  1.0 ml  of the
surrogate  spiking solution  to  the vial.   For a  solvent
miscible  sample,   dilute  the   sample   to  10   ml  with  the
appropriate  solvent  after  adding  1.0  ml  of  the  surrogate
spiking solution.   Cap and shake for 2 min.

      NOTE: Sections   7.3.3.2.1  and   7.3.3.2.2  must   be
            performed  rapidly   and  without interruption  to
            avoid  loss  of  volatile  organics.    These  steps
            must  be  performed  in  a  laboratory free  from
            solvent fumes.

      7.3.3.2.3   Pipet approximately 1 ml of the extract into
a  GC   vial  for  storage,  using a  disposable   pipet.    The
remainder may be  discarded.  Transfer  approximately 1  ml of
reagent methanol  to a separate GC vial  for use   as the method
blank for each set of samples.   These extracts  may be stored
at 4°C in the dark,  prior to analysis.

      7.3.3.2.4   The  GC  system  should  be set up   as  in
Section 7.0 of the specific determinative method. This should
be  done  prior to  the  addition of  the methanol  extract  to
organic-free reagent water.

      7.3.3.2.5   Table 2 can be used to determine the volume
of methanol extract to add to the 5 ml of organic-free reagent
water for analysis.   If  a screening procedure  was followed,
use the estimated concentration to determine the appropriate
volume.  Otherwise,  estimate the  concentration range  of the
sample from the low-concentration  analysis to  determine the
appropriate volume.  If  the sample  was submitted as a high-
concentration sample, start with 100 /xL.  All  dilutions must
keep  the  response  of  the  major   constituents (previously
saturated peaks)  in the upper half  of the linear range of the
curve.

      7.3.3.2.6   Remove the plunger from  a  5.0 ml Luerlock
type  syringe  equipped  with  a  syringe  valve and fill  until
overflowing with  organic-free  reagent  water.    Replace the
plunger and compress the water  to  vent trapped air.   Adjust
the volume  to  4.9 ml.    Pull the  plunger back  to 5.0  ml  to
allow volume  for  the addition of the  sample extract  and  of

                5030A - 11                        Revision 1
                                                   July 1992

-------
                  standards. Add 10 jil_ of internal  standard solution.  Also add
                  the volume of methanol extract determined  in Section 7.3.3.2.5
                  and a volume of methanol solvent to total  100  juL (excluding
                  methanol  in  standards).

                        7.3.3.2.7   Attach the syringe-syringe valve assembly to
                  the syringe valve on  the  purging device.   Open  the syringe
                  valve and inject the water/methanol sample  into  the purging
                  chamber.

                        7.3.3.2.8   Proceed with the analysis as outlined in the
                  specific  determinative method.  Analyze all reagent blanks on
                  the  same instrument  as that  used for  the  samples.    The
                  standards and blanks should also contain  100 jiL  of methanol
                  to simulate  the sample conditions.

                        7.3.3.2.9   For a matrix spike in the high-concentration
                  sediment/soil  samples,  add 8.0  ml of methanol,  1.0 ml  of
                  surrogate spike solution and  1.0 ml of matrix spike solution.
                  Add a  100 /zL  aliquot of this  extract  to 5 ml of  water for
                  purging  (as  per Section 7.3.3.2.6).

      7.4   Sample analysis:

            7.4.1 The samples  prepared by this method may  be analyzed by Methods
      8010, 8015, 8020, 8021,  8030,  8240,  and 8260.  Refer to these methods for
      appropriate analysis  conditions.


8.0   QUALITY CONTROL

      8.1   Refer to  Chapter  One for specific  quality control  procedures and
Method 3500 for sample preparation procedures.

      8.2   Before processing any samples,  the analyst should demonstrate through
the  analysis of  a  calibration  blank  that  all  glassware  and reagents are
interference free.  Each  time a set of samples is extracted,  or there is a change
in reagents, a method blank should be processed as a safeguard against chronic
laboratory contamination.  The blanks  should be carried through all  stages of
the sample preparation and measurement.

      8.3   Standard quality assurance practices should be used with  this method.
Field duplicates should be collected to validate the precision of the sampling
technique.  Laboratory replicates should be  analyzed to validate the precision
of the analysis.  Spiked samples  should  be carried  through  all stages of sample
preparation and measurement; they should be analyzed to validate the sensitivity
and accuracy of the analysis.   If the spiked  samples do not  indicate sufficient
sensitivity  to detect  <  1 jig/g  of the  analytes  in  the  sample,   then the
sensitivity  of  the  instrument  should  be increased,  or  the  sample  should be
subjected to additional cleanup.
                                  5030A - 12                        Revision  1
                                                                     July  1992

-------
9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods for performance data.


10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.
                                  5030A - 13                        Revision 1
                                                                     July 1992

-------
                                    TABLE  1
                      PURGE-AND-TRAP OPERATING  PARAMETERS
                                           Analysis Method
                             8010
                8015
              8020/8021
                8030
Purge gas
Purge gas flow rate
 (mL/min)
Purge time (min)
Purge temperature (°C)
Desorb temperature (°C)
Backflush inert gas flow
  (mL/min)
Desorb time (min)
Nitrogen or  Nitrogen or
  Heli urn

    40
11.0 ± 0.1
  Ambient
   180

  20-60
    4
  Helium
    20
15.0 ± 0.1
  85 + 2
   180

  20-60
   1.5
Nitrogen or  Nitrogen or
  Helium       Helium
    40
11.0 ± 0.1
  Ambient
   180

  20-60
    4
    20
15.0 + 0.1
  85 ± 2
   180

  20-60
   1.5
                                   5030A  -  14
                                           Revision  1
                                            July  1992

-------
                                    TABLE  2
            QUANTITY OF METHANOL EXTRACT REQUIRED FOR ANALYSIS OF
                      HIGH-CONCENTRATION SOILS/SEDIMENTS

                Approximate                         Volume of
            Concentration Range                  Methanol Extract8

              500-10,000 M9/kg                         100 /xL
            1,000-20,000 /ig/kg                           so ML
            5,000-100,000 jug/kg                          10 ^i
           25,000-500,000 jug/kg               100 /zL  of  1/50 dilution  b


Calculate appropriate dilution factor for concentrations exceeding this table.

"The volume of methanol  added to 5 ml of water being purged should be kept
constant.  Therefore, add to the 5 mL syringe whatever volume of methanol is
necessary to maintain a volume of 100 /iL added to the syringe.

Dilute an aliquot of the methanol  extract and then  take 100 jiL for analysis.
                                  5030A - 15                        Revision 1
                                                                     July 1992

-------
                          Figure 1
                      Purging Chamber
         OPTIONAL
         FOAM TRAP
* loch 0. D.
     10 mm Ot«t* Frit
     Moftwm Awmitv
                               Inlti Jt Inch 0. 0.
Swnplf Intet

2-W«v Svnnoj
17 cm. 20 Gfluoj Syrinot
6 mm 0. 0. Rubber Stptum


  -10 mm 0.0.
                                       Inch 0. D.
                      1/16 Inch 0 0.
                             Stit:
                                                      Ftow Control
                           5030A  -  16
                                  Revision  1
                                   July  1992

-------
                              Figure  2
       Trap Packing and Construction for Method  8010
           Packing Proctduri
                           Construction
 GlauWOOl

 Activatad
 Charcoal
 Grada 15
 Silica Gal
3%OV-1
GlauWoei
7.7cm
   Tanax    7.7 cm
8mm
           711/Foot
           Ratiitanca
           Wira Wrapped
           Solid
           (Doubla Layar)
               711/Foot
               Rtiistanc*
               Wira Wrapptd
               Solid
               (Singla Lavar)
                         8cm
Comprauion
Fining Nut
and Farrulai
                                                    Tharmoeouplt/
                                                    Controllar
                                                    Santer
        Elactronic
        Tamparatura
        Control and
        Pyromatar
  Tubing 25 cm
  0.105 In. 1.0.
  0.125 In. 0.0.
  Stainittt Staal
                  Trap Inlat
                             5030A  -  17
                                                            Revision  1
                                                             July 1992

-------
                                    Figure  3
        Trap Packing  and  Construction for Methods  8020  and  8030
            Packing Procedure
                                        Construction
Glass Wool    5mm  %
   Tenax   23 cm
3%OV-1   1cm';

Glass Wool    5 mm
                                                         Compression Fining Nut
                                                         and Ferrules

                                                           14 Ft. ?n/Foot Resistance
                                                           Wire Wrapped Solid
                                                              Thermocouple/Controller Sensor
                                                                   Electronic
                                                                   Temperature
                                                                   Control and
                                                                   Pyrometer
                                                            Tubing 25 cm
                                                            0.105 In. 1.0.
                                                            0.125 In. O.D.
                                                            Stainless Steel
                   Trap Inlet
                                     5030A -  18
                                                                              Revision 1
                                                                               July  1992

-------
                           Figure 4
                    Purge-and-Trap System
                       Purge-Sorb Mode
              For  Method 8010,  8020, and 8030
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
UOWO INJECTON PORTS

      COLUMN OVEN

               CONFIRMATORY COLUMN

              TO DETECTOR
PURGE OAS
FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
                                                ANALYTICAL COLUMN
                              OPTIONAL WORT COLUMN
                              SELECTION VALVE
                                PURGING
                                DEVICE
             NOTE
             ALL LINES BETWEEN TRAP
             AND GC SHOULD BE HEATED
             TO80-C
                           5030A -  19
                               Revision 1
                                July  1992

-------
                            Figure  5
                     Purge-and-Trap System
                          Desorb Mode
               For  Method 8010, 8020,  and 8030
CARPJCftGAS
FLOW CONTROL
PRESSURE
REGULATOR
UOUtO INJECTION PORTS

     COLUMN OVEN
                              OPTIONAL 4PORT COLUMN
                              SELECTION VALVE
PURGE GAS
ROW CONTROL
13X MOLECULAR
SIEVE FILTER
                                                CONFIRMATORY COLUMN
                                               TO DETECTOR
                                                ANALYTICAL COLUMN
                                 PURGING
                                 DEVICE
             NOTE:
             ALL LINES BETWEEN TRAP
             AND GC SHOULD BE HEATED
             TO SOX.
                           5030A - 20
                                Revision  1
                                 July  1992

-------
    METHOD 5030A
   PURGE-AND-TRAP
Start


7 1 Calibrate
CC system


712 Assemble
purge-and-trap
device and
condi tion t rap


712 Connect
to gas
chroma tograph
I
7 1 3 P
fin
solut i
repare
ons






714 Corry out
purge-and- trap
analysis




715 Calculate
response or
calibration factors
for each analyte
(Method 8000)


7 1 6
Calculate
averafe RF
for each
compound

/ '


/


     5030A -  21
Revision 1
 July 1992

-------
                                      METHOD  5030A
                                         continued
    7331
    Prepare
  samples and
   set-up CC
    sys tem
   73314
 Heigh sample
  into tared
    device
                         /
               Low concentration

                 Soil/sediment
            High concentration

              Soil/sediment
               water  samples  and
               water-miscible  wastes
                        	I
   73315
 Neigh another
  sample and
  determine %
  dry weight
 7 3 3  1  6  Add
spiked  reagent
water,  connect
   device to
    sys tem
   73317
   Heat and
 purge sample
/      •      /
  731  Screen
samples  prior to
 purge-and- trap
analysis,  dilute
 water miscible
    liquids
  731 Prepare
   sample and
  purg-and-trap
     device
                              7317
                              Dilute
                             purgeable
                             samples
   7 3 1 8 Add
  surrogate and
internal spiking
  solutions (if
   indicated)
                            7332  Add
                             methano 1
                            ex t ract  to
                           reagent  water
                           for ana lysis
                                                     7 3 3 2 Set
                                                   up CC sys tem
73326  Fill
 syringe  wi th
reagent wa ter ,
 vent air  and
ad j us t vo 1 ume
                           7 3 3 2 6 Add
                             internal
                           s tanda rd and
                             methano 1
                             extract
    Ana 1 y ze
 acco rd ing t o
 determinative
    me thod
                     7319
                  Inject sample
                  into chamber ,
                     purge
                    7 3 1 11
                   Desorb trap
                    into CC
                                                                       7  3  1  13
                                                                      Recondi tion
                                                                       trap and
                                                                       start  gas
                                                                         flow
                  7 3 1 13 Stop
                  gas flow and
                  cool trap for
                  next sample
                                                                        Analyze
                                                                     acco r ding  to
                                                                     determinative
                                                                        method
                                           5030A -  22
                                                                Revision 1
                                                                  July  1992

-------
                                 METHOD 5040A

  ANALYSIS OF SORBENT CARTRIDGES FROM VOLATILE ORGANIC SAMPLING TRAIN (VOST):
                GAS CHROMATOGRAPHY/MASS SPECTROMETRY  TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 5040 was formerly Method 3720 in the Second Edition of this
manual.

      1.2   This method covers the determination of volatile principal organic
hazardous constituents  (POHCs),  collected  on  Tenax and Tenax/charcoal sorbent
cartridges  using  a volatile  organic  sampling train,  VOST  (1).    Much  of the
description for purge-and-trap GC/MS analysis is described in Method 8240 of this
chapter.  Because the majority of gas  streams  sampled  using VOST will contain a
high concentration of water, the analytical  method  is  based on the quantitative
thermal desorption of volatile POHCs from the Tenax  and Tenax/charcoal traps and
analysis by purge-and-trap GC/MS.  For  the purposes of definition, volatile POHCs
are those POHCs with boiling  points less than  100°C.

      1.3   This  method  is  applicable to  the analysis  of  Tenax  and  Tenax/
charcoal cartridges used to collect volatile POHCs  from wet stack gas effluents
from hazardous waste incinerators.

      1.4   The sensitivity of the analytical method  for a particular volatile
POHC depends on the level  of interferences and the presence of detectable levels
of  volatile  POHCs  in  blanks.    The  desired  target  detection  limit of  the
analytical method is  0.1 ng/L  (20 ng on a single pair  of traps) for a particular
volatile POHC  desorbed  from either a  single  pair  of  Tenax  and  Tenax/charcoal
cartridges or  by  thermal  desorption of  up  to  six pairs of traps  onto a single
pair of Tenax  and Tenax/charcoal traps.  The resulting single pair of traps is
then thermally desorbed and analyzed by purge-and-trap GC/MS.

      1.5   This  method  is  recommended  for  use   only   by  experienced  mass
spectroscopists or under  the  close supervision of such qualified persons.


2.0   SUMMARY OF METHOD

      2.1   A  schematic diagram of the analytical  system is shown in Figure 1.
The contents of the sorbent cartridges are  spiked with an internal standard and
thermally desorbed for 10 min  at 180°C with  organic-free nitrogen or helium gas
(at a  flow  rate of 40  mL/min),  bubbled through  5  ml of  organic-free reagent
water,  and  trapped  on  an analytical  adsorbent   trap.    After  the  10  min.
desorption,  the analytical adsorbent trap is rapidly  heated to 180°C,  with the
carrier gas flow reversed so that the  effluent  flow from the analytical trap is
directed  into  the GC/MS.   The  volatile  POHCs are   separated by  temperature
programmed gas chromatography  and detected by low-resolution mass spectrometry.
The concentrations of volatile POHCs are calculated using the internal  standard
technique.
                                   5040A  -  1                         Revision 1
                                                                September 1994

-------
3.0   INTERFERENCES

      3.1   Refer to Methods 3500 and 8240.


4.0   APPARATUS AND MATERIALS

      4.1   Thermal desorption unit:

            4.1.1 The   thermal   desorption   unit   (for   Inside/Inside   VOST
      cartridges,  use  Supelco  "clamshell"  heater;   for   Inside/Outside  VOST
      cartridges,  user-fabricated   unit  is  required)  should  be  capable  of
      thermally desorbing the sorbent resin tubes.  It  should also be capable of
      heating the  tubes to  180 +  10°C with  flow of  organic-free  nitrogen  or
      helium through the tubes.

      4.2   Purge-and-trap unit:

            4.2.1 The purge-and-trap unit consists of  three separate pieces of
      equipment:   the sample purger,  trap,  and the  desorber.   It  should  be
      capable  of meeting  all requirements  of  Method  5030  for   analysis  of
      purgeable organic compounds from water.

      4.3   GC/MS system:  As described in Method 8240.


5.0   REAGENTS

      5.1   Organic-free reagent  water.   All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.2   Methanol, CH3OH - Pesticide grade, or equivalent.

      5.3   Analytical trap reagents:

            5.3.1 2,6-Diphenylene oxide polymer:  Tenax (60/80 mesh), chromato-
      graphic grade or equivalent.

            5.3.2 Methyl silicone packing:  3% OV-1 on  Chromosorb W  (60/80 mesh)
      or equivalent.

            5.3.3 Silica  gel:   Davison  Chemical  (35/00  mesh),  Grade  15,  or
      equivalent.

            5.3.4 Charcoal:  Petroleum-based (SKC Lot  104 or equivalent).

      5.4   Stock standard solution:

            5.4.1 Stock standard  solutions will be prepared from pure standard
      materials or purchased  as certified solutions.  The stock standards should
      be prepared in  methanol using assayed  liquids or  gases,  as  appropriate.
      Because of the toxicity of some of the organohalides,  primary dilutions of
      these materials should  be prepared  in a hood.  A  NIOSH/MESA approved toxic


                                   5040A  -  2                         Revision 1
                                                                September 1994

-------
      gas respirator should be used when the analyst  handles high concentrations
      of such materials.

            5.4.2 Fresh stock standards should be prepared weekly for volatile
      POHCs with boiling points of <35°C.  All other standards must be replaced
      monthly, or sooner if comparison with check standards indicates a problem.

      5.5   Secondary dilution standards:

            5.5.1 Using  stock   standard   solutions,   prepare,   in  methanol,
      secondary  dilution  standards  that  contain the  compounds  of  interest,
      either singly or mixed together.   The  secondary dilution standards should
      be prepared at concentrations such that the desorbed  calibration standards
      will bracket the working range of the analytical  system.

      5.6   4-Bromofluorobenzene (BFB) standard:

            5.6.1 Prepare  a 25 ng//xL solution of BFB in methanol.

      5.7   Deuterated benzene:

            5.7.1 Prepare  a 25 ng/^uL solution of benzene-d6 in methanol.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Refer to Method 0030, Chapter Ten.

      6.2   Sample trains  obtained from the VOST should be analyzed within 2-6
weeks of sample collection.


7.0   PROCEDURE

      7.1   Assembly of PTD device:

            7.1.1 Assemble  a purge-and-trap desorption device (PTD) that meets
      all the requirements  of Method 5030 (refer to Figure 1).

            7.1.2 Connect  the  thermal  desorption  device  to  the  PTD  device.
      Calibrate the PTD-GC/MS system using the internal standard technique.

      7.2   Internal standard calibration procedure:

            7.2.1 This approach  requires the  use of deuterated  benzene as the
      internal standard  for these  analyses.  Other  internal  standards may be
      proposed  for  use  in certain  situations.   The  important  criteria  for
      choosing a  particular compound as an  internal  standard are that  it be
      similar in analytical behavior to the compounds  of  interest and that it
      can be  demonstrated  that  the  measurement of  the  internal  standard be
      unaffected by method or matrix  interferences.   Other internal  standards
      that have been used  are ethylbenzene-d10 and,  l-2-dichloroethane-d4.  One
      adds 50 ng of BFB  to all  sorbent  cartridges  (in  addition  to one or more


                                   5040A -  3                        Revision 1
                                                                September 1994

-------
internal  standards)  to  provide  continuous   monitoring   of  the  GC/MS
performance relative to BFB.

      7.2.2 Prepare  calibration   standards  at  a   minimum   of  three
concentration levels for each analyte of interest.

      7.2.3 The  calibration  standards are  prepared  by spiking  a blank
Tenax or Tenax/charcoal  trap with a methanolic solution of the calibration
standards (including 50 ng  of  the  internal  standard, such as deuterated
benzene), using  the  flash  evaporation technique.   The flash evaporation
technique requires  filling the  needle  of  a  5.0  /jL  syringe  with clean
methanol and drawing air  into  the  syringe to the 1.0  fj,L  mark.   This is
followed by drawing  a  methanolic  solution  of the calibration standards
(containing 25 Aig/juL of the internal  standard)  to the 2.0 /iL mark.  The
glass  traps  should  be  attached  to  the   injection  port  of  a  gas
chromatograph while  maintaining  the  injector temperature  at 160°C.   The
carrier  gas  flow through  the traps  should  be  maintained at  about 50
mL/min.

      7.2.4 After directing the gas flow through the trap,  the contents of
the  syringe  should  be slowly  expelled  through  the  gas  chromatograph
injection port over  about  15 sec.  After 25 sec have  elapsed, the gas  flow
through  the trap should be shut off, the syringe removed,  and the  trap
analyzed by the  PTD-GC/MS  procedure  outlined  in Method 8240.   The total
flow of gas  through  the traps during  addition  of calibration standards to
blank cartridges, or internal standards to sample cartridges, should be 25
ml or less.

      7.2.5 Analyze  each calibration  standard for both Tenax  and Tenax/
charcoal cartridges  according to Section  7.3.   Tabulate the area response
of the characteristic  ions  of  each analyte  against  the concentration of
the  internal  standard  and calculate  the  response factor  (RF)  for  each
compound, using  Equation 1.

      RF = AsC,s/AisCs                                                  (1)

where:

      As  =       Area  of  the  characteristic ion for  the  analyte to be
                  measured.

      P(.   =        Area  of  the  characteristic   ion   for  the  internal
       •is
                  standard.

      CIS  =       Amount (ng) of the internal standard.

      Cs   =      Amount   (ng)  of  the  volatile  POHC  in  calibration
                  standard.

      If the RF value over  the working range is a constant (<10% RSD), the
RF can  be  assumed to be invariant,  and the average RF  can  be used for
calculations.    Alternatively,   the  results  can   be  used  to plot  a
calibration curve of response ratios, AS/AIS versus  RF.


                             5040A - 4                          Revision 1
                                                          September 1994

-------
            7.2.6 The working calibration curve or RF must be verified on each
      working  day  by  the  measurement  of  one  or  more  of the  calibration
      standards.  If the response varies  by more  than  ±25% for  any analyte, a
      new calibration standard must be prepared and analyzed for that analyte.

      7.3   The schematic of  the  PTD-GC/MS  system is shown  in  Figure  1.   The
sample cartridge  is placed in  the  thermal  desorption apparatus  (for  Inside/
Inside VOST cartridges,  use  Supelco  "clamshell" heater; for Inside/Outside VOST
cartridges, user fabricated  unit is required)  and desorbed in the purge-and-trap
system by heating to 180°C for 10 min at a flow  rate of 40 mL/min.  The desorbed
components pass into the bottom of the  water column, are purged from the water,
and collected  on  the  analytical  adsorbent trap.   After  the  10  min desorption
period, the compounds are desorbed from the analytical  adsorbent trap into the
GC/MS system according to the procedures described in Method 8240.

      7.4   Qualitative analysis

            7.4.1 The qualitative  identification of compounds determined by this
      method is based on retention  time,  and on comparison of the sample mass
      spectrum,  after  background correction,  with characteristic  ions  in  a
      reference mass spectrum.  The reference mass  spectrum must be generated by
      the laboratory using  the conditions of this  method.   The  characteristic
      ions from the  reference  mass spectrum are defined to  be the three ions of
      greatest relative intensity,  or  any ions over  30%  relative  intensity if
      less than  three  such  ions  occur  in the  reference  spectrum.   Compounds
      should be identified as present when the  criteria below are  met.

                  7.4.1.1     The intensities of the characteristic  ions  of a
            compound maximize  in the same scan or within one scan of each other.
            Selection of a peak by a data system target compound search routine,
            where   the   search   is   based  on  the   presence  of   a   target
            chromatographic  peak containing   ions  specific  for  the  target
            compound at a compound-specific retention time, will be accepted as
            meeting  this criterion.

                  7.4.1.2     The  RRT  of  the  sample  component  is   within
            + 0.06 RRT units of the  RRT of the  standard component.

                  7.4.1.3     The relative  intensities  of  the  characteristic
            ions agree within 30% of the relative  intensities of these  ions in
            the reference spectrum.   (Example:  For an ion with an  abundance of
            50% in  the  reference spectrum, the  corresponding abundance  in  a
            sample spectrum can range between  20%  and 80%.)

                  7.4.1.4     Structural isomers that produce very similar mass
            spectra  should  be  identified as individual  isomers  if  they  have
            sufficiently different GC retention times.  Sufficient GC resolution
            is achieved if the  height of the valley between two isomer peaks is
            less than  25% of  the sum  of  the  two peak  heights.    Otherwise,
            structural  isomers are identified  as  isomeric pairs.

                  7.4.1.5     Identification  is hampered when sample components
            are  not resolved  chromatographically and   produce  mass  spectra


                                  5040A  - 5                         Revision 1
                                                                September 1994

-------
      containing  ions  contributed  by more than  one analyte.   When gas
      chromatographic  peaks  obviously  represent more  than one  sample
      component  (i.e.,  a broadened  peak with  shoulder(s)  or  a  valley
      between  two or  more maxima),  appropriate  selection of  analyte
      spectra  and background  spectra  is  important.    Examination  of
      extracted ion current profiles of  appropriate  ions  can aid  in the
      selection  of  spectra,  and   in   qualitative  identification  of
      compounds.  When  analytes  coelute  (i.e.,  only one chromatographic
      peak is apparent), the  identification criteria  can be met, but each
      analyte  spectrum  will  contain  extraneous  ions contributed  by the
      coeluting compound.

      7.4.2 For  samples containing components  not   associated  with the
calibration standards,  a  library search  may  be made for  the purpose of
tentative  identification.   The  necessity  to  perform  this  type  of
identification will be determined by the type  of analyses being conducted.
Guidelines for making tentative  identification are:

      (1)   Relative intensities of major ions in the reference spectrum
(ions >  10%  of the most  abundant  ion)  should be present  in the  sample
spectrum.

      (2)   The relative intensities of the major ions should  agree within
+ 20%.   (Example:   For an ion  with an abundance of  50% in the standard
spectrum, the  corresponding sample ion abundance must  be  between  30 and
70%).

      (3)   Molecular ions present  in  the reference spectrum  should be
present in the sample spectrum.

      (4)   Ions present in the  sample spectrum but not in the reference
spectrum  should be  reviewed  for  possible  background  contamination or
presence of coeluting compounds.

      (5)   Ions present in the  reference spectrum but not in the sample
spectrum  should be  reviewed  for possible  subtraction from  the  sample
spectrum because  of  background contamination or coeluting  peaks.   Data
system   library   reduction   programs   can    sometimes   create   these
discrepancies.

      Computer  generated  library  search  routines   should   not  use
normalization  routines  that  would misrepresent  the library  or unknown
spectra when compared to each other.  Only after  visual  comparison of the
sample  with   the  nearest  library  searches  will   the  mass  spectral
interpretation specialist assign a tentative identification.

7.5   Quantitative analysis

      7.5.1 When   an   analyte   has   been   qualitatively   identified,
quantitation should be based  on the integrated  abundance from the EICP of
the primary  characteristic  ion chosen for that  analyte.   If the sample
produces an interference for the primary  characteristic ion, a secondary
characteristic  ion should be used.
                             5040A -  6                        Revision 1
                                                          September 1994

-------
                  7.5.1.1     Using the internal standard calibration procedure,
            the amount of  analyte  in  the sample cartridge is calculated using
            the response factor (RF) determined in Section 7.2.5  and  Equation 2.

                  Amount of POHC  =  AsCis/Ai8RF                              (2)

            where:

                  As =  Area of  the  characteristic ion  for  the analyte to be
                        measured.

                  Ais =  Area  for  the  characteristic   ion  of  the  internal
                        standard.

                  Cis =  Amount (ng) of  internal standard.

                  7.5.1.2     The  choice   of   methods  for  evaluating  data
            collected using  VOST for incinerator trial  burns  is a regulatory
            decision.  The procedures used  extensively by one user are outlined
            below.

                  7.5.1.3     The  total  amount  of  the  POHCs  of  interest
            collected on a pair of traps should be summed.

                  7.5.1.4     The observation of high concentrations of POHCs of
            interest   in  blank   cartridges   indicates   possible  residual
            contamination of the  sorbent cartridges prior to shipment to  and use
            at  the  site.   Data  that fall  in  this category (especially data
            indicating high concentrations of POHCs in  blank sorbent  cartridges)
            should be qualified with regard to validity,  and blank  data should
            be reported separately.  The applicability of data of this  type to
            the determination  of  ORE  is  a regulatory  decision.    Continued
            observation  of  high  concentrations  of  POHCs  in   blank   sorbent
            cartridges  indicates  that  procedures  for   cleanup,   monitoring,
            shipment, and storage of  sorbent cartridges by a  particular  user be
            investigated to eliminate this problem.

                  7.5.1.5     If any internal standard recoveries fall  outside
            the control  limits  established in Section 8.4, data for all analytes
            determined  for  that  cartridge(s)  must   be  qualified with  the
            observation.
8.0   QUALITY CONTROL

      8.1   Refer to  Chapter  One for specific quality  control  procedures and
Method 0030 for sample preparation procedures.

      8.2   Each laboratory that uses this method  is required to operate a formal
quality control  program.  The minimum requirements of this  program consist of an
initial demonstration of laboratory capability and the analysis of blank Tenax
and  Tenax/charcoal  cartridges  spiked with  the  analytes of  interest.    The
laboratory is required to maintain performance records to  define the quality of


                                   5040A  -  7                         Revision 1
                                                                September 1994

-------
data that  are generated.   Ongoing performance  checks  must be  compared  with
established performance criteria to determine if results  are within the expected
precision and accuracy limits of the method.

            8.2.1 Before performing any analyses, the analyst must demonstrate
      the  ability  to  generate acceptable  precision and  accuracy with  this
      method.  This ability is established as described in Section 7.2.

            8.2.2 The  laboratory   must  spike   all  Tenax  and  Tenax/charcoal
      cartridges with the internal  standard(s) to monitor continuing laboratory
      performance.   This procedure is described in Section 7.2.

      8.3   To  establish  the  ability  to  generate  acceptable  accuracy  and
precision,  the analyst must spike blank Tenax and Tenax/charcoal cartridges with
the analytes of interest at two concentrations in the working range.

            8.3.1 The  average  response factor  (Rf)  and  the  standard deviation
      (s) for each must be calculated.

            8.3.2 The  average recovery and standard deviation must fall within
      the expected range for determination of volatile POHCs using this method.
      The expected range for recovery of volatile POHCs using this method is 50-
      150%.

      8.4   The  analyst must  calculate  method  performance  criteria  for  the
internal standard(s).

            8.4.1 Calculate  upper   and   lower  control  limits   for  method
      performances   using   the  average   area   response   (A)   and  standard
      deviation(s) for internal standard:

            Upper Control Limit (UCL) = A + 3s
            Lower Control Limit (LCL) = A - 3s

            The UCL  and  LCL  can  be used  to  construct control  charts  that are
      useful  in  observing  trends  in performance.   The  control  limits must be
      replaced by method performance criteria as they  become available from the
      U.S.  EPA.

      8.5   The laboratory is required  to  spike all  sample cartridges (Tenax and
Tenax/charcoal) with  internal  standard.

      8.6   Each day,  the  analyst must demonstrate  through  analysis  of blank
Tenax  and  Tenax/charcoal   cartridges  and  organic-free  reagent   water  that
interferences from the analytical  system  are under control.

      8.7   The  daily GC/MS performance  tests  required  for  this  method  are
described  in Method 8240.
9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods for performance data.
                                   5040A - 8                        Revision 1
                                                                September 1994

-------
10.0  REFERENCES

1.     Protocol  for  Collection  and  Analysis  of Volatile  POHC's Using  VOST.
      EPA/600/8-84-007, March 1984.

2.     Validation  of the  Volatile  Organic  Sampling  Train  (VOST)  Protocol.
      Volumes I and II.  EPA/600/4-86-014a, January 1986.
                                  5040A - 9                         Revision 1
                                                                September 1994

-------

«» •
», •
•
*••
t
t
i
Ik


t:
G
                              C CTl
                              O —i
                              •r- 0,
                              > .Q
                              O) £
                              Q£ 0)
                               •(->
                                Q.
                                0)
                               C/}
                         i
                         |
                         I
                         O   O
                         •o   S
                         X   S
z
4
IS


K^IJ 1 H
i?Sr«j 2; S 6
i * -Z

8f*
V

                         £
                         s.

-------
                                          METHOD 5040A
ANALYSIS  OF  SORBENT  CARTRIDGES FROM  VOLATILE  ORGANIC SAMPLING TRAIN
                   GAS  CHROMATOGRAPHY/MASS SPECTROMETRY TECHNIQUE
                                                                (VOST)
 (    Start     j
 7.1.1 Assemble
 purge and trap
   desorption
     device.
  7.1.2 Connect
     thermal
    desorption
     device;
  calib. system.
   7.2.1 Select
     internal
    standard.
   7.2.3 Prepare
     calibration
  standards using
  flash evaporat.
     technique.
    7.2.4 Direct
     gas flow
   through traps.
  7.2.4 Expel
  contents of
syringe through
  GC injection
     port.
 7.2.4 Analyze
 trap by P-T-D
    GC/MS
   procedure.
  7.2.5 Analyze
   •ach calib.
  standard for
 both cartridges
   (aee 7.3).
 7.2.5 Tabulate
 area response
 and calculate
response factor.
   7.2.6 Verify
    response
   factor each
      day.
    7.3 Place
     sample
   cartridge in
desorp. apparatus;
  desorb in P-T.
    7.3 Desorb
    into GC/MS
     system.
       7.4.1
    Qualitatively
      identify
  volatile POHCs.
      7.5.1 Use
       primary
     characteristic
       ion for
     quantitation.
       7.6.1.1
      Calculate
  amount of analyte
      in sample.
  7.5.1.3 Sum
amount of POHCs
  of interest for
  each pair of
     traps.
 7.5.1.4 Examine
 blank* data for
 •igns of residual
 contamination.
 7.5.1.5 Compare
     int. std.
   recoveries to
    Section 8.4
   control limits.
                                                                                     Stop
                                           5040A -  11
                                                              Revision  1
                                                         September  1994

-------
                                  METHOD  5041

       PROTOCOL  FOR ANALYSIS OF  SORBENT CARTRIDGES  FROM  VOLATILE  ORGANIC
          SAMPLING  TRAIN  (VOST):  WIDE-BORE  CAPILLARY  COLUMN  TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   This method  describes  the analysis of  volatile  principal  organic
hazardous  constituents  (POHCs)  collected  from the  stack  gas effluents  of
hazardous waste incinerators using the VOST methodology (1).  For  a comprehensive
description of  the  VOST  sampling methodology see Method  0030.   The following
compounds may be determined by this method:
      Compound Name
CAS No.'
Acetone
Acrylonitrile
Benzene
Bromodi chl oromethane
Bromoformb
Bromomethane0
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chl orodi bromomethane
Chloroethanec
Chloroform
Chl oromethane0
Di bromomethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans- 1,2-Di chl oroethene
1 , 2-Di chl oropropane
ci s- 1 ,3-Di chl oropropene
trans-l,3-Dichloropropene
Ethyl benzene6
lodomethane
Methylene chloride
Styreneb
l,l,2,2-Tetrachloroethaneb
Tetrachloroethene
Toluene
67-64-1
107-13-1
71-43-2
75-27-4
75-25-2
74-83-9
75-15-0
56-23-5
108-90-7
124-48-1
75-00-3
67-66-3
74-87-3
74-95-3
75-35-3
107-06-2
75-35-4
156-60-5
78-87-5
10061-01-5
10061-02-6
100-41-4
74-88-4
75-09-2
100-42-5
79-34-5
127-18-4
108-88-3
                                                                   (continued)
                                   5041  -  1
                Revision 0
            September 1994

-------
      Compound Name                                 CAS No.a


      1,1,1-Trichloroethane                         71-55-6
      1,1,2-Trichloroethane                         79-00-5
      Trichloroethene                               79-01-6
      Trichlorofluoromethane                        75-69-4
      l,2,3-Trichloropropaneb                       96-18-4
      Vinyl chloride0                               75-01-4
      Xylenesb


      8   Chemical  Abstract Services  Registry Number.

      b   Boiling  point  of this compound  is above 132°C.   Method  0030  is not
appropriate for quantitative sampling of this analyte.

      0   Boiling point of  this compound is below  30°C. Special precautions must
be taken when sampling  for  this analyte  by  Method 0030.  Refer to Sec. 1.3 for
discussion.

      1.2   This method  is most successfully applied to the analysis  of non-polar
organic compounds with boiling points between 30°C and 100°C.  Data are applied
to the calculation of destruction and removal efficiency  (ORE), with limitations
discussed below.

      1.3   This method may be applied to analysis of many compounds which boil
above 100°C,  but Method  0030 is always inappropriate for collection  of compounds
with boiling points above  132°C. All  target  analytes with  boiling  points greater
than 132°C are so noted in the target analyte list presented in Sec. 1.1.  Use
of Method  0030  for collection of  compounds boiling between  100°C  and 132°C is
often  possible,   and  must  be  decided  based on  case by  case  inspection  of
information  such   as  sampling method  collection efficiency,  tube desorption
efficiency, and analytical method  precision and  bias.  An organic compound with
a boiling  point below  30°C  may break through the sorbent under the conditions
used for  sample  collection.  Quantitative  values obtained  for compounds with
boiling points below 30°C  must be qualified, since the  value obtained represents
a minimum value  for the compound if breakthrough  has occurred.  In  certain cases,
additional QC measures may have been  taken during sampling very low  boilers with
Method  0030.     This   information   should  be  considered  during  the  data
interpretation stage.

      When Method 5041 is  used  for survey analyses, values for compounds boiling
above 132°C may  be reported  and qualified since the quantity obtained represents
a minimum  value for  the compound.  These minimum  values should not be used for
trial burn ORE calculations or to prove  insignificant risk.

      1.4   The VOST analytical methodology can  be used to quantitate volatile
organic  compounds  that are  insoluble  or  slightly  soluble  in  water.   When
volatile,  water soluble  compounds  are  included  in the  VOST organic compound
analyte list, quantitation limits can be expected to be approximately  ten  times

                                    5041  - 2                        Revision  0
                                                                 September 1994

-------
higher than quantitation limits for water insoluble  compounds  (if the compounds
can be recovered at all)  because the purging efficiency from water (and possibly
from Tenax-GC®)  is  poor.

      1.5   Overall  sensitivity  of the method is dependent  upon  the level of
interferences  encountered  in  the  sample  and   the   presence  of  detectable
concentrations of volatile POHCs in blanks.  The  target detection limit of this
method is 0.1 pg/m3 (ng/L) of flue gas, to permit calculation  of a ORE equal to
or greater  than 99.99% for volatile  POHCs which may  be  present  in the waste
stream at 100 ppm.  The upper end of the range of applicability of this method
is limited by the dynamic range of the analytical instrumentation, the overall
loading  of  organic compounds  on  the  exposed  tubes,   and  breakthrough  of the
volatile POHCs on the sorbent traps used to collect the  sample.  Table 1 presents
retention times  and characteristic  ions  for volatile compounds which  can be
determined by this method.  Table 2 presents method detection limits  for a range
of volatile compounds analyzed by the wide-bore VOST methodology.

      1.6   The wide-bore VOST analytical methodology  is restricted to use by,
or under the supervision of, analysts experienced in the use of sorbent media,
purge-and-trap systems,  and gas chromatograph/mass spectrometers, and skilled in
the interpretation of mass spectra and their use as a  quantitative tool.


2.0  SUMMARY OF METHOD

      2.1   The sorbent tubes  are thermally desorbed by heating and purging with
organic-free helium.   The gaseous effluent  from  the  tubes is bubbled through
pre-purged organic-free reagent water and trapped on an analytical  sorbent trap
in a purge-and-trap unit (Figure 2).  After desorption, the analytical sorbent
trap is heated rapidly and the gas  flow from the  analytical trap is directed to
the head of a wide-bore column  under subambient conditions.  The volatile organic
compounds  desorbed  from  the  analytical   trap  are  separated by  temperature
programmed  high resolution gas  chromatography  and  detected  by  continuously
scanning low resolution mass spectrometry (Figure  3).  Concentrations  of volatile
organic compounds are calculated from a multi-point calibration curve, using the
method of response factors.


3.0   INTERFERENCES

      3.1   Sorbent tubes which are to be analyzed for volatile  organic compounds
can be  contaminated by  diffusion of volatile  organic  compounds  (particularly
Freon® refrigerants and common organic  solvents)  through the external container
(even through a Teflon® lined  screw cap on  a  glass container) and the Swagelok®
sorbent tube caps during shipment  and  storage.   The  sorbent tubes  can also be
contaminated if organic  solvents  are  present  in the analytical laboratory.  The
use of blanks  is  essential  to assess the  extent  of  any  contamination.   Field
blanks must be prepared and taken to the field.  The end caps of the tubes are
removed for the period of time required  to exchange  two  pairs of  traps  on the
VOST sampling apparatus.  The tubes are recapped and  shipped and handled exactly
as the actual  field  samples are shipped and handled.   At least  one pair of field
blanks is included with each six pairs of sample cartridges collected.
                                   5041 - 3                         Revision 0
                                                                September 1994

-------
      3.2   At  least  one  pair  of  blank  cartridges  (one  Tenax-GC®,  one
Tenax-GC®/charcoal) shall be included with shipment of cartridges to a hazardous
waste incinerator site as trip blanks.   These trip blanks will  be treated like
field blanks except that the end  caps will not be removed during storage at the
site.  This pair of traps will  be  analyzed  to  monitor potential  contamination
which may occur during storage and shipment.

      3.3   Analytical   system   blanks   are  required   to   demonstrate   that
contamination  of  the  purge-and-trap   unit and  the  gas  chromatograph/mass
spectrometer has not occurred or  that, in the event of analysis of sorbent tubes
with very high  concentrations of organic compounds,  no  compound  carryover is
occurring.  Tenax® from the  same  preparation batch as the Tenax® used for field
sampling should be used  in  the  preparation  of  the method (laboratory)  blanks.
A sufficient number of  cleaned  Tenax®  tubes from the  same  batch  as  the  field
samples should be reserved in the laboratory for use as blanks.

      3.4   Cross contamination can occur whenever low-concentration samples are
analyzed after high-concentration  samples,  or  when  several  high-concentration
samples are analyzed  sequentially.   When an unusually concentrated  sample is
analyzed, this analysis should be followed by a method blank to establish that
the  analytical  system  is  free  of contamination.    If  analysis  of  a  blank
demonstrates that the  system is contaminated, an additional bake cycle should be
used.  If the analytical system  is still contaminated after additional  baking,
routine system maintenance  should  be performed:  the analytical  trap  should be
changed and  conditioned,  routine  column  maintenance should be  performed (or
replacement of the  column  and  conditioning of the new column,  if necessary), and
bakeout of the  ion  source  (or  cleaning of the ion source and rods, if required).
After system maintenance has been performed,  analysis  of  a blank is required to
demonstrate that the cleanliness of the system is acceptable.

      3.5   Impurities in the purge gas and from organic compounds out-gassing
in tubing account  for the majority of  contamination problems.   The analytical
system must be demonstrated to be free from contamination under the conditions
of the  analysis by analyzing two sets of clean, blank sorbent tubes with organic-
free reagent purge water as  system  blanks.   The  analytical system is acceptably
clean when these two  sets of blank  tubes  show values  for  the analytes which are
within one  standard  deviation  of  the  normal  system  blank.   Use of  plastic
coatings,  non-Teflon®  thread   sealants,  or  flow  controllers  with  rubber
components should be avoided.

      3.6   VOST tubes are handled in the laboratory to spike standards and to
position the tubes within the desorption  apparatus.   When  sorbent  media are
handled in  the  laboratory atmosphere,  contamination  is  possible  if  there are
organic solvents in use anywhere in the laboratory.  It is therefore necessary
to make daily use of system blanks to monitor the cleanliness  of the sorbents and
the absence of  contamination from the analytical system.  A single set of system
blank  tubes  shall be  exposed  to  normal  laboratory handling  procedures and
analyzed as a sample.   This  sample should be within one  standard deviation of
normal  VOST  tube  blanks to  demonstrate  lack of contamination of  the  sorbent
media.

      3.7   If the emission  source has a high concentration of non-target organic
compounds (for example,  hydrocarbons at  concentrations  of hundreds of ppm), the


                                   5041 - 4                         Revision 0
                                                                September 1994

-------
presence of these non-target compounds will  interfere with the performance of the
VOST  analytical  methodology.   If one  or  more of  the compounds  of interest
saturates  the  chromatographic  and   mass   spectrometric  instrumentation,  no
quantitative calculations can be made and the tubes which have been sampled under
the same conditions will  yield no valid data for any of  the saturated compounds.
In the presence of  a very high organic loading, even if the compounds of interest
are not saturated,  the  instrumentation is so saturated that the linear range has
been surpassed. When instrument saturation  occurs, it is possible that compounds
of  interest cannot  even  be  identified correctly  because  a saturated  mass
spectrometer may  mis-assign  masses.    Even  if compounds of interest  can  be
identified,  accurate  quantitative   calculations  are  impossible  at  detector
saturation.  No determination  can be made  at  detector  saturation,  even if the
target compound itself  is not saturated.  At detector saturation, a negative bias
will be encountered in analytical measurements and no  accurate calculation can
be made for the Destruction and Removal  Efficiency if  analytical  values may be
biased negatively.

      3.8   The recoveries of  the surrogate compounds,  which are spiked on the
VOST  tubes  immediately before analysis, should  be  monitored carefully as  an
overall indicator  of the  performance of the methodology.  Since  the matrix of
stack emissions is  so variable, only  a general  guideline for  recovery of  50-150%
can be used for surrogates.  The  analyst cannot use the surrogate recoveries as
a guide for correction  of compound recoveries.   The surrogates are valuable only
as a general indicator of  correct operation of the methodology.   If surrogates
are not observed or if  recovery of one or more of  the surrogates is outside the
50-150% range,  the  VOST methodology is not operating correctly. The cause of the
failure  in  the methodology  is not  obvious.   The  matrix  of stack  emissions
contains large  amounts of water,  may be highly acidic, and  may  contain large
amounts  of target  and non-target organic compounds.    Chemical  and   surface
interactions may be occurring on the tubes.  If recoveries of surrogate compounds
are  extremely  low or  surrogate compounds  cannot even be   identified  in  the
analytical process, then failure to observe an analyte may or may not imply that
the compound of interest has been  removed from the emissions  with a high degree
of efficiency (that is, the Destruction  and  Removal Efficiency for that  analyte
is high).


4.0   APPARATUS AND MATERIALS

      4.1   Tube desorption apparatus: Acceptable performance of the methodology
requires:   1)  temperature  regulation to ensure  that  tube  temperature  during
desorption is regulated to  180°C + 10°;   2)  good  contact between  tubes and the
heating apparatus  to ensure that  the sorbent bed is thoroughly  and  uniformly
heated  to  facilitate  desorption  of organic  compounds;     and  3)  gas-tight
connections to  the  ends of the tubes to ensure flow of desorption gas through the
tubes without leakage during the heating/desorption process.   A simple clamshell
heater which will  hold  tubes  which  are 3/4"  in  outer diameter will  perform
acceptably as a desorption apparatus.

      4.2   Purge-and-trap device:  The  purge-and-trap  device consists of three
separate pieces of equipment:  a sample  purge vessel, an analytical  trap, and a
desorber.  Complete devices are commercially  available from a  variety of sources,
or the separate components may be assembled.  The cartridge  thermal  desorption


                                   5041  - 5                         Revision 0
                                                                September 1994

-------
apparatus  is  connected  to  the  sample  purge  vessel  by  1/8"  Teflon®  tubing
(unheated transfer line).  The tubing which connects the desorption chamber to
the sample purge vessel should be as short as is practical.

            4.2.1 The sample purge vessel  is required to hold 5 ml of organic-
      free reagent water, through which the gaseous effluent  from the VOST tubes
      is routed.  The water column  should  be  at  least 3  cm  deep.   The gaseous
      headspace between the water  column  and the analytical trap  must  have a
      total volume of  less  than 15 ml.   The purge gas must pass  through the
      water column as finely divided bubbles with a diameter of less than 3 mm
      at  the   origin.    The  sample purger  shown  in  Figure  4  meets  these
      requirements.  Alternate sample purging  vessels may be used if equivalent
      performance is demonstrated.

            4.2.2 The  analytical  trap must  be  at  least 25 cm  and have  an
      internal diameter of  at  least 0.105  in.  The analytical trap must contain
      the following components:

            2,6-diphenylene oxide polymer:      60/80 mesh, chromatograph grade
                                                (Tenax-GC®,  or equivalent)

            methyl silicone packing:            OV-1 (3%) on Chromosorb-W 60/80
                                                mesh, or equivalent

            silica gel:                         35/60 mesh,  Davison grade  15 or
                                                equivalent

            coconut charcoal:                    prepare  from Barneby  Cheney,
                                                CA-580-26,   or  equivalent,  by
                                                crushing   through   26   mesh
                                                screen.

            The  proportions  are:   1/3  Tenax-GC®,  1/3  silica  gel,  and  1/3
      charcoal,  with  approximately  1.0  cm  of methyl  silicone packing.  The
      analytical trap should be conditioned for four  hours at 180°C with gas flow
      (10 mL/min)  prior  to  use in sample  analysis.   During  conditioning, the
      effluent of the trap should not be vented to the analytical  column.  The
      thermal  desorption apparatus is connected to the injection system of the
      mass spectrometer by a transfer line which is heated to 100°C.

            4.2.3 The desorber must be capable of rapidly heating the analytical
      trap to 180°C for desorption.  The polymer section of the trap should not
      exceed 180°C, and the remaining sections should not exceed 220°C,  during
      bake-out mode.

      4.3   Gas chromatograph/mass spectrometer/data system:

            4.3.1 Gas  chromatograph:   An  analytical system complete  with  a
      temperature  programmable  oven with  sub-ambient temperature  capabilities
      and all  required  accessories, including  syringes, analytical  columns, and
      gases.
                                   5041 - 6                         Revision 0
                                                                September 1994

-------
            4.3.2 Chromatographic column:  30 m x  0.53 mm  ID  wide-bore fused
      silica capillary column, 3 /urn film thickness, DB-624  or equivalent.

            4.3.3 Mass spectrometer:   capable of scanning from 35-260 amu every
      second or  less,  using 70  eV  (nominal) electron energy  in  the electron
      ionization  mode  and  producing  a mass  spectrum that  meets all  of  the
      criteria in Table 3 when 50 ng of 4-bromofluorobenzene (BFB) is injected
      into the water in the purge vessel.

            4.3.4 Gas  chromatograph/mass   spectrometer   interface:   Any  gas
      chromatograph  to  mass  spectrometer  interface  that  gives  acceptable
      calibration points at 50  ng or less per injection of each of the analytes,
      and achieves the performance criteria for  4-bromofluorobenzene shown in
      Table 3,  may be used.   If a glass jet separator is used with the wide-bore
      column, a helium  make-up  flow of approximately 15 ml,  introduced after the
      end of the  column  and  prior  to the  entrance  of  the effluent  to  the
      separator, will be required for optimum performance.

            4.3.5 Data system:   A  computer  system  that allows  the continuous
      acquisition  and  storage  on machine  readable  media  of all  mass  spectra
      obtained throughout  the  duration  of  the Chromatographic  program must be
      interfaced to the mass spectrometer.  The computer must have software that
      allows searching  any gas  chromatographic/mass  spectrometric data file for
      ions of a specified mass and plotting such ion abundances versus time or
      scan number.   This  type  of plot  is  defined as  an Extracted Ion Current
      Profile  (EICP).     Software  must  also  be   available that  allows  the
      integration of the  ion abundances in any  EICP between specified time or
      scan number limits.   The most recent version  of the EPA/NIST Mass Spectral
      Library should also be available.

      4.4   Wrenches:  9/16", 1/2",  7/16", and 5/16".

      4.5   Teflon® tubing:  1/8" diameter.

      4.6   Syringes: 25 ^uL  syringes  (2),  10 /j,l syringes (2).

      4.7   Fittings:  1/4" nuts, 1/8" nuts, 1/16" nuts, 1/4" to  1/8" union, 1/4"
to 1/4" union,  1/4" to  1/16" union.

      4.8   Adjustable  stand  to raise  the level  of  the  desorption  unit,  if
required.

      4.9   Volumetric  flasks:   5 ml,  class A with ground  glass stopper.

      4.10  Injector port or equivalent, heated to  180°C  for loading standards
onto VOST tubes prior to analysis.

      4.11  Vials:  2 ml, with Teflon® lined screw caps or crimp tops.

      4.12  Syringe:   5 ml, gas-tight with shutoff valve.
                                   5041 - 7                         Revision 0
                                                                September 1994

-------
5.0   REAGENTS

      5.1   Reagent grade chemicals  shall be used in all tests. Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications
of the Committee on Analytical Reagents  of the American Chemical Society, where
such specifications are available. Other grades may  be used,  provided it  is first
ascertained that the reagent  is  of  sufficiently  high  purity  to permit  its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

            5.2.1 It is advisable to maintain the stock of organic-free reagent
      water generated for use in  the purge-and-trap apparatus with a continuous
      stream of inert gas bubbled through the water. Continuous bubbling of the
      inert gas maintains a positive pressure of  inert gas above the water as a
      safeguard against contamination.

      5.3   Methanol,  CH3OH.    Pesticide quality   or  equivalent.   To  avoid
contamination with  other laboratory solvents,   it  is  advisable to  maintain a
separate stock of methanol for the preparation of standards for VOST analysis and
to regulate the use of this methanol very carefully.

      5.4   Stock  standard solutions  - Can be prepared  from pure  standard
materials or can be purchased as  certified solutions.  Stock standard solutions
must be prepared in high purity methanol.  All  preparation of standards should
take place in a hood, both  to avoid contamination  and  to ensure  safety of the
analyst preparing the standards.

            5.4.1 Place about 4 ml of high purity methanol in a 5 ml volumetric
      flask.  Allow the flask to  stand,  unstoppered, for  about 10 min, or until
      all alcohol wetted surfaces have dried.

                  5.4.1.1     Add appropriate volumes of neat liquid chemicals
            or certified solutions,  using a syringe of the appropriate volume.
            Liquid which  is added to the volumetric  flask must  fall  directly
            into the alcohol without contacting  the neck  of the flask.  Gaseous
            standards  can  be purchased as  methanol  solutions   from  several
            commercial  vendors.

                  5.4.1.2     Dilute  to volume with  high  purity  methanol,
            stopper, and then mix by inverting the  flask  several times.  Calcu-
            late concentration by the  dilution  of  certified  solutions  or neat
            chemicals.

            5.4.2 Transfer  the  stock standard  solution  into a Teflon® sealed
      screw  cap  bottle.    An amber  bottle  may  be used. Store,  with  minimal
      headspace, at -10°C to -20°C,  and  protect  from light.

            5.4.3 Prepare fresh standards every two months for gases. Reactive
      compounds such as  styrene  may need to  be  prepared  more frequently.   All
      other standards must be replaced after six months, or sooner  if comparison
      with check standards  indicates a problem.


                                    5041  - 8                         Revision 0
                                                                September 1994

-------
      5.5   Secondary  dilution  standards:    Using  stock  standard  solutions,
prepare,  in high  purity  methanol,  secondary dilution standards containing the
compounds  of  interest,  either  singly  or mixed together.   Secondary dilution
standards must be stored  with minimal headspace and should be checked  frequently
for  signs  of  degradation  or  evaporation,  especially just prior  to preparing
calibration standards from them.

      5.6   Surrogate  standards:  The  recommended  surrogates are  toluene-d8,
4-bromofluorobenzene, and l,2-dichloroethane-d4. Other compounds may be  used as
surrogate compounds, depending upon the requirements of the analysis.  Surrogate
compounds are selected to span  the elution range of the compounds  of interest.
Isotopically labeled compounds  are selected to preclude the observation of the
same compounds in the stack emissions.   More than  one  surrogate is  used  so that
surrogate measurements can still be made even if analytical interferences with
one or more of the surrogate compounds are encountered.  However, at least three
surrogate compounds  should  be used to monitor the performance of the  methodology.
A stock surrogate compound solution in high purity methanol should be prepared
as described  in Sec. 5.4,  and  a surrogate  standard spiking solution should be
prepared  from the  stock  at  a  concentration  of  250  jug/10 ml in  high purity
methanol.  Each pair of VOST tubes (or each individual VOST tube,   if the tubes
are  analyzed  separately)  must  be spiked with 10  juL  of  the  surrogate  spiking
solution prior to GC/MS analysis.

      5.7   Internal  standards:    The  recommended  internal  standards  are
bromochloromethane,  1,4-difluorobenzene, and chlorobenzene-d5.  Other compounds
may be used as internal standards as long as  they have retention times  similar
to the  compounds  being analyzed by GC/MS.    The  internal  standards should be
distributed through  the chromatographic elution range.  Prepare  internal standard
stock  and secondary  dilution  standards  in  high purity  methanol  using  the
procedures described  in  Sees.  5.2 and  5.3.   The  secondary  dilution standard
should be prepared at a concentration of 25 mg/L of each of the  internal standard
compounds.  Addition of  10 fj,l  of this  internal  standard solution  to each pair
of VOST tubes  (or to each VOST tube, if  the tubes are  analyzed individually) is
the equivalent of 250 ng total.

      5.8   4-Bromofluorobenzene (BFB) standard:  A standard solution containing
25 ng/juL of BFB in  high purity  methanol should be prepared for use as a tuning
standard.

      5.9   Calibration standards:  Calibration standards at  a minimum of five
concentrations will  be required  from the secondary dilution of stock standards
(see Sees.  5.2  and  5.3).   A range of  concentrations for calibration  can be
obtained by use  of  different  volumes  of a  50 mg/L  methanol  solution  of  the
calibration  standards.     One   of  the  concentrations  used   should be  at  a
concentration  near, but  above, the method  detection  limit.   The remaining
concentrations should correspond to the expected  range of concentrations found
in field samples  but should not  exceed the linear range of the GC/MS analytical
system (a typical  range for a calibration would be  10,  50,  100, 350, and 500 ng,
for  example).    Each  calibration  standard  should  contain  each   analyte  for
detection by this method.  Store calibration  standards  for one week only in a
vial  with no headspace.
                                   5041 - 9                         Revision 0
                                                                September 1994

-------
      5.10  Great care must be taken to maintain the integrity of all standard
solutions.  All standards of  volatile  compounds  in  methanol  must be stored at
-10° to  -20°C  in amber bottles with Teflon® lined screw caps or crimp tops.  In
addition, careful  attention must be paid to the use of syringes designated for
a specific purpose or for use with only a single standard solution since cross
contamination of volatile organic standards can occurs very readily.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See Method 0030 for the VOST Sampling Methodology.

      6.2   VOST samples are collected  on paired cartridges.   The first of the
pair of  sorbent cartridges  is  packed  with approximately  1.6 g  of  Tenax-GC®
resin.   The second cartridge of  the pair is packed with Tenax-GC® and petroleum
based charcoal  (3:1  by  volume;  approximately  1 g of each).   In  sampling, the
emissions gas stream passes through the Tenax-GC® layer first and then through
the charcoal  layer.   The Tenax-GC® is cleaned and reused;  charcoal  is not reused
when tubes are prepared.  Sorbent is cleaned and the tubes are  packed.  The tubes
are desorbed and subjected  to a  blank  check prior  to  being  sent  to the field.
When the tubes are used for sampling (see Figure 5 for a schematic  diagram of the
Volatile  Organic  Sampling  Train (VOST)), cooling  water is circulated  to the
condensers and the temperature of the cooling water is maintained  near 0°C.  The
end caps  of  the sorbent cartridges are placed in a clean, screw capped glass
container during sample collection.

      6.3   After  the  apparatus   is  leak   checked,   sample  collection  is
accomplished by opening the valve to the first condenser, turning on the pump,
and sampling at a  rate of  1  liter/min for  20 minutes.  The volume of sample for
any pair of traps  should not exceed 20 liters.   An alternative  set of conditions
for sample collection requires sampling  at  a reduced flow rate,  where the overall
volume of sample collected is 5 liters  at a rate of 0.25 L/min for 20 minutes.
The 20 minute period is required for collecting an integrated sample.

      6.4   Following collection  of 20  liters of  sample,  the  train  is  leak
checked a second time at the highest pressure drop encountered  during the run to
minimize the chance of vacuum desorption of organics from the Tenax®.

      6.5   The train is returned  to atmospheric pressure  and the two sorbent
cartridges are removed.  The end caps are replaced and the cartridges are placed
in a suitable environment  for storage and  transport  until analysis.  The sample
is considered invalid if the leak test  does not meet specifications.

      6.6   A new pair  of  cartridges  is  placed in  the VOST,  the VOST is leak
checked, and the sample collection  process is  repeated until six  pairs of traps
have been exposed.

      6.7   All sample  cartridges  are kept  in  coolers on  cold  packs  after
exposure and during shipment.  Upon  receipt at the laboratory,  the  cartridges are
stored in a refrigerator at 4°C  until analysis.
                                   5041  -  10                         Revision 0
                                                                September 1994

-------
7.0   PROCEDURE

      7.1    Recommended    operating   conditions   for   cartridge    desorber,
purge-and-trap unit, and gas chromatograph/mass spectrometer using the wide-bore
column are:
      Cartridge Desorption Oven
      Desorb Temperature
      Desorb Time
      Desorption Gas Flow
      Desorption/Carrier Gas

      Purge-and-Trap Concentrator
      Analytical Trap Desorption Flow
      Purge Temperature
      Purge Time
      Analytical Trap Desorb Temperature
      Analytical Trap Desorb Time

      Gas Chromatoqraph
      Column
      Carrier Gas Flow
      Makeup Gas Flow
      Injector Temperature
      Transfer Oven Temperature
      Initial Temperature
      Initial Hold Time
      Program Rate
      Final Temperature
      Final Hold Time

      Mass Spectrometer
      Manifold Temperature
      Scan Rate
      Mass Range
      Electron Energy
      Source Temperature
180°C
11 minutes
40 mL/min
Helium, Grade 5.0
2.5 mL/min helium
Ambient
11 minutes
180°C
5 minutes
DB-624,  0.53  mm  ID x  30 m  thick
film (3 jum) fused silica capillary,
or equivalent
15 mL/min
15 mL/min
200°C
240°C
5°C
2 minutes
6°C/min
240°C
1 minute, or until  elution  ceases
105°C
1 sec/cycle
35-260 amu
70 eV (nominal)
According    to
specifications
manufacturer's
      7.2   Each GC/MS  system  must be hardware tuned to  meet  the criteria in
Table 3 for a  50 ng injection of 4-bromofluorobenzene (2  juL injection  of the BFB
standard  solution  into  the water  of the purge  vessel).   No  analyses  may be
initiated until the criteria presented in Table 3 are met.

      7.3   Assemble a  purge-and-trap device  that  meets the specifications in
Method 5030.  Condition  the analytical trap overnight at  180°C in the purge mode,
with an inert  gas  flow of at  least  20 mL/min.   Prior to  use each day,  condition
the trap for  10 minutes by backflushing at 180°C, with the column  at  220°C.

      7.4   Connect the purge-and-trap device to a gas chromatograph.
                                   5041  -  11
                         Revision 0
                     September  1994

-------
      7.5   Assemble  a  VOST  tube  desorption   apparatus   which   meets  the
requirements of Sec. 4.1.

      7.6   Connect the  VOST  tube  desorption apparatus to  the  purge-and-trap
unit.

      7.7   Calibrate the instrument using the internal  standard procedure, with
standards  and  calibration  compounds   spiked  onto  cleaned  VOST  tubes  for
calibration.

            7.7.1 Compounds in methanolic solution are  spiked onto  VOST tubes
      using the flash  evaporation technique.  To perform flash evaporation, the
      injector of  a gas  chromatograph  or an equivalent piece of  equipment is
      required.

                  7.7.1.1     Prepare a syringe with  the appropriate volume of
            methanolic standard solution (either surrogates,  internal standards,
            or calibration compounds).

                  7.7.1.2     With  the injector port  heated to 180°C,  and with
            an inert gas flow of 10 mL/min  through the injector  port,  connect
            the paired VOST tubes  (connected as in Figure  1, with  gas  flow in
            the same direction as the sampling gas flow) to the  injector port;
            tighten with  a wrench so  that  there  is  no leakage  of gas.   If
            separate  tubes  are   being  analyzed, an   individual   Tenax®  or
            Tenax®/charcoal tube is connected to the  injector.

                  7.7.1.3     After directing  the gas  flow through the VOST
            tubes, slowly inject the first standard solution over a period of 25
            seconds.  Wait  for 5  sec before withdrawing the syringe  from the
            injector port.

                  7.7.1.4     Inject  a  second standard  (if required)  over  a
            period  of  25 seconds  and  wait  for 5 sec  before withdrawing  the
            syringe from the injector port.

                  7.7.1.5     Repeat  the sequence  above as required until  all of
            the necessary compounds are spiked onto the VOST tubes.

                  7.7.1.6     Wait  for  30 seconds, with gas  flow, after the last
            spike before disconnecting  the tubes.  The total time the tubes are
            connected to the injector port with gas flow should  not exceed 2.5
            minutes.   Total  gas  flow  through  the tubes  during the  spiking
            process should not exceed 25 ml to prevent break  through of adsorbed
            compounds  during  the  spiking  process.    To allow  more time  for
            connecting and disconnecting tubes, an on/off valve may be installed
            in the  gas  line to the injector  port so  that  gas is  not  flowing
            through the tubes during  the connection/disconnection process.

      7.8   Prepare the purge-and-trap unit with  5 ml  of organic-free reagent
water in the purge vessel.
                                   5041  -  12                         Revision 0
                                                                September 1994

-------
      7.9   Connect the paired VOST tubes to the gas lines in the tube desorption
unit.  The tubes must be connected so that the gas flow during desorption will
be opposite to the  flow of  gas  during  sampling:  i.e.,  the tube desorption gas
passes through the charcoal  portion of the tube first.   An on/off valve may be
installed in the gas line leading to the tube desorption unit in order to prevent
flow of gas through the tubes during the connection process.

      7.10  Initiate tube desorption/purge and heating  of  the  VOST tubes in the
desorption apparatus.

      7.11  Set the oven of  the  gas  chromatograph to subambient temperatures by
cooling with liquid nitrogen.

      7.12  Prepare the GC/MS system for data acquisition.

      7.13  At the conclusion of the tube/water purge time,  attach the analytical
trap to the gas chromatograph,  adjust  the purge-and-trap  device  to the desorb
mode,  and  initiate  the gas   chromatographic  program   and  the  GC/MS  data
acquisition.  Concurrently,   introduce   the   trapped   materials   to  the  gas
chromatographic column  by rapidly heating the analytical  trap  to 180°C while
backflushing the  trap  with  inert gas at  2.5  mL/min  for  5  min.   Initiate the
program for the gas chromatograph and simultaneously initiate data acquisition
on the GC/MS system.

      7.14  While  the   analytical  trap   is   being  desorbed  into  the  gas
chromatograph, empty the purging  vessel.  Wash the  purging  vessel with a minimum
of two  5  ml flushes  of organic-free reagent  water  (or  methanol  followed by
organic-free  reagent  water) to  avoid  carryover  of  analytes  into  subsequent
analyses.

      7.15  After the sample has been desorbed, recondition  the analytical trap
by employing a bake cycle on the purge-and-trap unit.   The analytical  trap may
be  baked   at  temperatures  up  to  220°C.    However,   extensive   use  of  high
temperatures  to  recondition  the trap  will   shorten  the   useful  life  of the
analytical trap.   After approximately  11  minutes,  terminate the  trap  bake and
cool  the trap to  ambient temperatures in  preparation for the next sample.  This
procedure is a convention for reasonable samples and should be adequate if the
concentration of contamination  does not  saturate the analytical system.  If the
organic compound concentration is so high  that the analytical system is saturated
beyond the point where even  extended system bakeout is  not sufficient to clean
the system, a more extensive system maintenance must be performed.  To perform
extensive system maintenance, the analytical trap is replaced and the new trap
is conditioned.   Maintenance is  performed on the GC column  by  removing at least
one foot  from  the front end  of the column.   If the  chromatography  does not
recover after column maintenance, the chromatographic  column must be replaced.
The ion source should  be baked  out and,  if the bakeout  is  not  sufficient to
restore mass spectrometric  peak  shape and  sensitivity,  the ion  source and the
quadrupole rods must be cleaned.

      7.16  Initial calibration  for the  analysis of VOST tubes: It is essential
that calibration  be performed in the mode in which analysis will  be performed.
If tubes are being analyzed  as pairs,  calibration standards should be  prepared
                                   5041  -  13                         Revision 0
                                                                September 1994

-------
on paired tubes.  If tubes are being analyzed  individually, a calibration should
be performed on individual Tenax® only tubes and Tenax®/charcoal tubes.

            7.16.1       Prepare the calibration  standards by spiking VOST tubes
      using the procedure described in Sec. 7.7.1.   Spike each pair  of VOST
      tubes  (or  each of  the individual  tubes)  immediately  before  analysis.
      Perform the calibration analyses in order from low concentration to high
      to minimize the compound carryover.  Add  5.0  ml of organic-free reagent
      water to the  purging vessel.   Initiate  tube desorb/purge according to the
      procedure described above.

            7.16.2       Tabulate the area response of the characteristic primary
      ions  (Table  1) against  concentration  for each  target compound,  each
      surrogate compound, and each internal standard.   The first criterion for
      quantitative   analysis is  correct  identification  of  compounds.    The
      compounds must elute  within  + 0.06 retention time units  of  the elution
      time of the standard analyzed on  the  same analytical system on the day of
      the analysis.   The analytes should be quantitated relative to the closest
      eluting  internal  standard, according  to  the scheme  shown  in  Table  4.
      Calculate response factors  (RF) for each compound relative to the internal
      standard  shown  in  Table   4.   The  internal   standard  selected  for  the
      calculation of  RF  is  the  internal  standard  that  has a  retention  time
      closest to the  compound being measured.  The RF is calculated as follows:

            RF = (Ax/Cis)/(Als/Cx)

      where:

            Ax  =  area  of  the   characteristic   ion   for  the  compound  being
                  measured.

            Ais =  area  of the  characteristic  ion  for  the specific  internal
                  standard.

            C,s =  concentration of the specific internal  standard.

            Cx  =  concentration of the compound being  measured.

            7.16.3       The average RF must be calculated for each compound. A
      system performance  check should  be made before  the  calibration curve is
      used.  Five compounds  (the System Performance Check Compounds,  or SPCCs)
      are checked for a minimum  average response  factor.   These compounds are
      chloromethane,  1,1-dichloroethane, bromoform,  1,1,2,2-tetrachloroethane,
      and chlorobenzene.  The minimum  acceptable  average RF for these compounds
      should be 0.300 (0.250 for  bromoform).  These compounds typically have RFs
      of 0.4  - 0.6,  and are used to check compound instability  and  check for
      degradation caused  by  contaminated lines or active  sites  in  the system.
      Examples of these  occurrences are:

                  7.16.3.1    Chloromethane:   This compound is the most likely
            compound to  be lost  if the purge flow is too fast.
                                   5041  -  14                         Revision 0
                                                                September 1994

-------
            7.16.3.2    Bromoform:  This  compound  is one of the compounds
      most likely to be purged very  poorly  if the  purge flow is too slow.
      Cold  spots  and/or  active  sites in  transfer lines  may  adversely
      affect response.  Response of the primary  quantitation ion (m/z 173)
      is directly affected by the tuning for 4-bromofluorobenzene at the
      ions of masses 174 and  176.  Increasing the ratio  of  ions 174 and 176
      to   mass   95   (the   base   peak   of   the  mass   spectrum   of
      bromofluorobenzene)  may improve bromoform response.

            7.16.3.3    1,1,2,2-Tetrachloroethane and 1,1-dichloroethane:
      These  compounds are degraded by  contaminated  transfer lines  in
      purge-and-trap  systems and/or active sites in trapping materials.

      7.16.4      Using the response factors from the initial calibration,
calculate  the  percent  relative  standard  deviation  (%RSD)   for  the
Calibration Check Compounds  (CCCs).
      %RSD  =  (SD/X) x 100
where:
      %RSD
      RF
      SD
   percent  relative  standard  deviation

   individual  RF  measurement

   mean  of  5 initial  RFs for a compound  (the  5 points
   over  the calibration  range)

   standard deviation  of average RFs  for a compound,
   where SD is calculated:
       SD =
             N
(RFi-RF)
   jn
      The %RSD  for  each  individual  CCC should be  less  than  30 percent.
This criterion must be met in order for the individual calibration to be
valid.  The CCCs are: 1,1-dichloroethene, chloroform, 1,2-dichloropropane,
toluene, ethyl benzene, and vinyl chloride.

7.17  Daily GC/MS Calibration

      7.17.1    Prior  to  the  analysis  of  samples,  purge  50  ng  of  the
4-bromofluorobenzene standard.  The resultant  mass  spectrum  for the  BFB
must meet  all  of the criteria  given  in  Table 3 before  sample analysis
begins.   These  criteria must  be  demonstrated  every  twelve  hours  of
operation.

      7.17.2    The initial calibration  curve (Sec. 7.16)  for each compound
of  interest  must be  checked and  verified once every  twelve  hours  of
analysis  time.     This   verification   is   accomplished  by  analyzing  a
                             5041  -  15
                                         Revision 0
                                     September 1994

-------
calibration  standard  that   is  at  a  concentration   near  the  midpoint
concentration for  the  working  range of the GC/MS  and checking  the SPCC
(Sec. 7.16.3) and CCC  (Sec.  7.16.4).

      7.17.3   System  Performance  Check  Compounds  (SPCCs):   A  system
performance check must be made  each  twelve hours of analysis. If the SPCC
criteria  are met,  a  comparison  of response  factors  is  made for  all
compounds.   This  is the  same  check that  is applied  during  the  initial
calibration.  If the minimum response factors are not  achieved, the system
must be evaluated, and corrective  action must be taken before analysis is
allowed to begin.   The  minimum response factor for volatile SPCCs is 0.300
(0.250  for  bromoform).    If  these minimum  response  factors  are  not
achieved,  some  possible  problems  may  be  degradation  of the  standard
mixture, contamination of the  injector port, contamination  at  the front
end  of  the  analytical   column,  and  active  sites   in the  column  or
chromatographic system.  If the problem is active sites at the front end
of the analytical  column, column maintenance (removal of approximately 1
foot from the front end of the column) may remedy the problem.

      7.17.4   Calibration Check Compounds:  After the system performance
check has  been  met, CCCs listed  in Sec.  7.16.4  are used to  check the
validity  of the   initial  calibration.  Calculate  the  percent  difference
using the following equation:

                        (RF,  - RFC)   x 100
      % Difference = 	
where:

      RFi =     average response factor from initial calibration

      RFC =     response factor from current calibration check standard.

      If the percent difference for any compound is greater than 20, the
laboratory should  consider  this  a warning limit.  Benzene,  toluene, and
styrene will  have problems with response factors if Tenax® decomposition
occurs  (either  as  a  result  of  sampling   exposure  or  temperature
degradation), since these  compounds are decomposition products of Tenax®.
If the  percent difference  for each  CCC  is less  than 25%,  the initial
calibration  is  assumed  to  be valid.    If  the   criterion  of  percent
difference less  than 25%  is not  met for  any  one  CCC,  corrective action
MUST be taken.   Problems similar to those  listed under SPCCs could affect
this criterion.   If a  source of the problem  cannot  be determined after
corrective action  is taken,  a new five-point  calibration  curve MUST be
generated.   The  criteria  for the  CCCs  MUST  be met  before quantitative
analysis can begin.

      7.17.5      Internal standard responses  and retention times in the
check calibration standard must be evaluated  immediately after or during
data acquisition.  If the retention time for  any internal standard changes
by more  than 30 seconds  from the last  check calibration (12  hr),  the
chromatographic system must be inspected for malfunctions and corrections


                             5041  - 16                         Revision 0
                                                          September 1994

-------
must be made,  as  required.   A factor which  may  influence the retention
times of the internal  standards  on  sample  tubes  is the level  of overall
organic compound loading on the  VOST  tubes.   If  the VOST tubes are very
highly  loaded  with either  a  single  compound  or  with  multiple  organic
compounds, retention times for standards and compounds of interest will be
affected.   If  the area  for  the  primary  ion  of  any  of  the  internal
standards changes by a factor of two (-50% to +100%) from the last daily
calibration check, the gas chromatograph and mass  spectrometer should be
inspected for malfunctions and corrections  must be made, as appropriate.
If the level of organic loading of  samples  is high,  areas for the primary
ions  of both  compounds of  interest  and  standards  will   be  adversely
affected.  Calibration  check standards should not  be subject to variation,
since the concentrations of  organic compounds on  these samples are set to
be  within  the  linear  range  of the  instrumentation.    If  instrument
malfunction has occurred, analyses of samples performed under conditions
of malfunction may be invalidated.

7.18  GC/MS Analysis of Samples

      7.18.1      Set  up  the  cartridge desorption unit,  purge-and-trap
unit, and GC/MS as described above.

      7.18.2      BFB tuning criteria and  daily  GC/MS calibration check
criteria must be met before analyzing samples.

      7.18.3      Adjust the  helium  purge gas  flow rate  (through  the
cartridges  and  purge  vessel)  to  approximately 40  mL/min.   Optimize the
flow rate to provide the best response for chloromethane  and bromoform, if
these compounds are analytes. A  flow  rate  which  is too high reduces the
recovery of chloromethane,  and an insufficient gas flow rate reduces the
recovery of bromoform.

      7.18.4      The first  analysis performed  after the tuning check and
the calibration or daily calibration check  is a method blank.  The method
blank consists  of clean VOST tubes (both Tenax® and Tenax®/charcoal) which
are spiked  with surrogate compounds  and  internal  standards according to
the procedure described in Sec.  7.7.1.  The tubes which are used for the
method blanks  should be from the same batch  of sorbent as the sorbent used
for the  field  samples.  After  the tubes  are  cleaned  and  prepared  for
shipment to the field, sufficient pairs of  tubes  should be retained from
the  same  batch in  the  laboratory  to provide  method blanks  during  the
analysis.

      7.18.5      The organic-free  reagent  water  for the purge vessel for
the analysis  of each of the  VOST samples  should be supplied  from  the
laboratory  inventory which is  maintained with constant bubbling of inert
gas to avoid contamination.

      7.18.6      If  the  analysis  of  a  pair   of VOST  tubes  has  a
concentration  of analytes that exceeds the  initial calibration range, no
reanalysis of  desorbed  VOST  tubes is possible.  An additional calibration
point can be added  to bracket the higher concentration encountered in the
samples so that the calibration  database encompasses six or more points.
                             5041  -  17                         Revision 0
                                                          September 1994

-------
Alternatively, the  data  may be  flagged  in the report  as  "extrapolated
beyond the upper range of the calibration." The use of the secondary ions
shown in Table 1  is  permissible  only  in the case of interference with the
primary quantitation  ion.   Use  of secondary ions  to  calculate compound
concentration  in  the case  of saturation  of the  primary  ion  is  not  an
acceptable procedure, since a negative bias of an unpredictable magnitude
is  introduced  into  the  quantitative data when  saturation  of  the  mass
spectrum of a compound is encountered.   If high organic loadings, either
of  a  single  compound or  of multiple compounds,  are  encountered,  it  is
vital that a  method blank be  analyzed prior to the analysis  of another
sample  to demonstrate  that no  compound  carryover   is  occurring.    If
concentrations of organic compounds are sufficiently high that carryover
problems are profound, extensive bakeout of the purge-and-trap unit will
be  required.   Complete replacement of  the contaminated analytical trap,
with the associated  requirement for conditioning the new trap, may  also be
required for VOST samples which  show excessive  concentrations of organic
compounds.  Other measures which  might be  required for decontamination of
the  analytical   system   include   bakeout of   the mass   spectrometer,
replacement of the filament of the mass spectrometer,  cleaning of the ion
source of the  mass  spectrometer, and/or  (depending  on the  nature of the
contamination) maintenance of the chromatographic column or replacement of
the  chromatographic   column,   with   the  associated  requirement   for
conditioning the new chromatographic column.

7.19  Data Interpretation

      7.19.1      Qualitative analysis:

            7.19.1.1    The  qualitative   identification   of   compounds
      determined  by this  method  is  based  on  retention  time,  and  on
      comparison of  the sample mass spectrum,  after background correction,
      with  characteristic  ions  in  a  reference  mass spectrum.    The
      reference mass spectrum must be generated  by the laboratory using
      the conditions  of  this method.  The characteristic  ions from the
      reference mass spectrum are defined to be  the three ions of greatest
      relative intensity, or any ions over 30% relative  intensity  if less
      than three  such  ions occur in the  reference  spectrum.   Compounds
      should be identified  as present when the  criteria below are met.

                  7.19.1.1.1  The intensities of the characteristic ions
            of a compound maximize  in the  same scan or within one  scan of
            each  other.    Selection  of  a peak  by  a  data  system target
            compound  search routine where  the  search  is  based  on the
            presence  of  a  target  chromatographic peak containing  ions
            specific  for  the  target compound  at  a   compound  specific
            retention time will  be accepted as  meeting this criterion.

                  7.19.1.1.2  The RRT of  the sample component  is + 0.06
            RRT units of the RRT of the standard component.
                  7.19.1.1.3  The    relative    intensities     of   the
            characteristic  ions  agree   within   30%   of  the  relative
            intensities  of these   ions   in   the  reference  spectrum.
            (Example:    For an   ion  with  an abundance of  50%  in the


                             5041  -  18                         Revision 0
                                                          September 1994

-------
      reference  spectrum,  the  corresponding  abundance in a sample
      spectrum can range between 20% and 80%.)

            7.19.1.1.4  Structural isomers that produce very similar
      mass  spectra  should  be identified as  individual  isomers if
      they  have  sufficiently  different   GC   retention  times.
      Sufficient  GC  resolution is  achieved  if the  height  of the
      valley between two isomer peaks is less than 25% of the sum of
      the  two  peak  heights.    Otherwise,  structural  isomers are
      identified as isomeric pairs.

            7.19.1.1.5  Identification  is  hampered  when  sample
      components  are  not resolved  chromatographically and produce
      mass  spectra  containing  ions contributed  by  more  than one
      analyte.   When gas chromatographic peaks obviously represent
      more  than  one  sample component  (i.e.,  a broadened peak with
      shoulder(s)  or  a   valley  between  two  or  more  maxima),
      appropriate  selection  of  analyte  spectra  and  background
      spectra is  important.   Examination  of  extracted ion current
      profiles  of appropriate  ions  can aid  in  the  selection of
      spectra,  and in  qualitative identification of compounds.  When
      analytes  coelute  (i.e.,  only one  chromatographic peak is
      apparent), the  identification criteria  can  be  met, but each
      analyte spectrum will contain extraneous ions contributed by
      the coeluting compound.

      7.19.1.2    For samples containing components not associated
with the calibration  standards,  a library search may be made for the
purpose of tentative identification.  The necessity to perform this
type of  identification will  be  determined by the  type of analyses
being conducted.   Guidelines for making  tentative  identification
are:

      (1)   Relative  intensities  of major  ions  in  the  reference
spectrum (ions  > 10% of  the most abundant ion) should be present in
the sample  spectrum.

      (2)   The relative intensities of  the major ions should agree
within + 20%.   (Example:  For an ion with an abundance of  50% in the
standard spectrum, the corresponding sample  ion  abundance must be
between 30  and 70%).

      (3)   Molecular ions  present  in the reference spectrum should
be present  in  the sample spectrum.

      (4)   Ions  present  in the  sample spectrum  but not  in  the
reference  spectrum  should be  reviewed for possible  background
contamination  or presence of coeluting compounds.

      (5)   Ions present in the reference spectrum but not  in the
sample spectrum should be reviewed for possible subtraction from the
sample spectrum  because  of background  contamination  or  coeluting
peaks.  Data system library reduction programs can sometimes create
these discrepancies.

                      5041 - 19                         Revision 0
                                                    September 1994

-------
      Computer generated  library search  routines should  not  use
normalization  routines that  would  misrepresent  the  library  or
unknown spectra  when  compared to each  other.    Only  after visual
comparison of sample with the nearest library searches will the mass
spectral    interpretation    specialist    assign   a    tentative
identification.

7.19.2      Quantitative analysis:

      7.19.2.1    When  a   compound  has   been   identified,   the
quantitative  analysis  of  that  compound  will   be based  on  the
integrated abundance from the extracted ion current profile of the
primary characteristic  ion  for that  compound  (Table  1).   In  the
event  that  there  is  interference  with  the primary  ion  so  that
quantitative measurements cannot  be made, a secondary  ion  may be
used.

      NOTE: Use  of   a  secondary   ion   to  perform  quantitative
            calculations in  the  event  of  the  saturation  of  the
            primary ion is not  an acceptable procedure  because of
            the unpredictable extent of the negative bias which is
            introduced.   Quantitative calculations are  performed
            using the  internal  standard technique.   The internal
            standard used to perform quantitative  calculations shall
            be the internal  standard nearest the retention time of
            a given analyte (see Table 4).

      7.19.2.2    Calculate the  amount  of each  identified analyte
from the VOST tubes as follows:

      Amount (ng)  =  (AsCis)/(AisRF)

where:

      As  = area of the  characteristic  ion  for  the analyte to be
            measured.

      Ais =  area of the characteristic ion of the  internal standard.

      Cis =  amount (ng) of the internal  standard.

      7.19.2.3    The  choice   of   methods  for   evaluating  data
collected using the VOST methodology for  incinerator trial burns is
a  regulatory decision.   Various  procedures are  used  to  decide
whether  blank  correction  should  be  performed  and  how  blank
correction should be performed.  Regulatory  agencies to which VOST
data are submitted also vary in their preferences  for data which are
or which are not blank corrected.

      7.19.2.4    The  total  amount  of   the  POHCs  of  interest
collected on a pair of traps should be summed.
                       5041  -  20                        Revision 0
                                                    September 1994

-------
                  7.19.2.5    The occurrence of high  concentrations of analytes
            on method blank cartridges indicates possible residual contamination
            of  sorbent  cartridges prior to  shipment and use  at  the sampling
            site.  Data with  high  associated blank values must be qualified with
            respect  to  validity,  and  all   blank   data  should  be  reported
            separately.   No blank  corrections should  be  made in  this case.
            Whether or not data  of this type can be applied to the determination
            of  destruction  and removal  efficiency   is  a  regulatory decision.
            Continued observation of  high  concentrations  of analytes on blank
            sorbent cartridges indicates  that procedures for cleanup  and quality
            control for  the  sampling  tubes are inadequate.   Corrective action
            MUST be  applied  to tube preparation  and monitoring  procedures to
            maintain  method  blank  concentrations  below detection  limits  for
            analytes.

                  7.19.2.6    Where applicable, an estimate of concentration for
            noncalibrated components  in  the  sample may  be  made.   The formulae
            for quantitative calculations  presented  above  should  be used with
            the following modifications:  The areas Ax and Ais should  be from the
            total  ion   chromatograms,   and  the  Response   Factor  for  the
            noncalibrated compound  should  be  assumed to be 1.    The  nearest
            eluting internal standard free from interferences in the total  ion
            chromatogram  should be  used  to determine the  concentration.   The
            concentration which is obtained should be reported indicating:  (1)
            that the value is an estimate;  and  (2) which internal  standard was
            used.

                  7.19.2.7    If any internal standard recoveries fall  outside
            the control  limits  established in  Sec. 8.4,  data for  all analytes
            determined  for  that   cartridge(s) must be  qualified  with  the
            observation.    Report  results  without   correction for  surrogate
            compound recovery data.   When  duplicates are  analyzed,  report  the
            data obtained with the sample  results.


8.0   QUALITY CONTROL

      8.1   Each laboratory  that  uses these methods  is required  to operate a
formal quality  control  program.   The  minimum quality control requirements  are
specified in Chapter One.  In addition,  this program should consist  of an initial
demonstration of laboratory  capability and  an  ongoing analysis of check samples
to evaluate and document data quality.  The laboratory must maintain records to
document the quality  of the  data  generated.   Ongoing data  quality  checks  are
compared with established performance  criteria to determine if the results of
analyses meet  the performance  characteristics of   the method.   When  sample
analyses indicate atypical method  performance,  a quality control check standard
(spiked method  blank)  must  be  analyzed to confirm that  the measurements were
performed in an in-control mode of instrument operation.

      8.2   Before processing  any  samples,  the  analyst  should  demonstrate,
through the analysis  of a method blank  (laboratory blank sorbent tubes,  reagent
water purge) that interferences from the analytical   system,  glassware,  sorbent
tube  preparation,  and  reagents  are under  control.   Each  time a  new  batch of


                                  5041 - 21                         Revision 0
                                                                September 1994

-------
sorbent tubes  is  analyzed,  a method blank should be  processed  as a safeguard
against chronic laboratory contamination.  Blank tubes which have been carried
through all the stages of  sorbent preparation and handling should be  used in the
analysis.

      8.3   The experience  of the  analyst performing  the GC/MS analyses  is
invaluable to the  success  of  the analytical methods.   Each day that the analysis
is  performed,  the  daily  calibration  check  standard  should  be  evaluated  to
determine  if the  chromatographic  and tube  desorption  systems  are operating
properly.  Questions that  should be asked are:  Do the peaks look  normal?  Is the
system response obtained comparable to the  response from  previous calibrations?
Careful examination of the chromatogram of the calibration standard can indicate
whether column maintenance is  required or  whether the column  is still  usable,
whether there  are leaks  in  the  system,  whether the  injector  septum requires
replacing, etc.  If changes are made to the system (such as change of a column),
a calibration check must be carried out and a  new multipoint calibration must be
generated.

      8.4   Required  instrument  quality  control  is  found  in   the following
sections:

            8.4.1 The mass spectrometer must be tuned to meet the specifications
      for 4-bromofluorobenzene in Sec. 7.2 (Table 3).

            8.4.2 An initial  calibration  of the tube desorption/purge-and-trap/
      GC/MS must be performed  as specified in Sec.  7.7.

            8,4.3 The GC/MS system must meet the SPCC criteria specified in Sec.
      7.16.3 and the CCC criteria in Sec.  7.16.4 each  twelve hours of  instrument
      operation.

      8.5   To  establish   the  ability  to generate   acceptable accuracy  and
precision, the analyst must perform the following operations.

            8.5.1 A quality  control  (QC)  check  sample concentrate is required
      containing  each  analyte  at  a  concentration  of  10 mg/L  in high  purity
      methanol.   The  QC  check sample concentrate may be prepared  from  pure
      standard materials  or purchased as certified solutions.    If the QC check
      sample  concentrate  is  prepared by  the  laboratory,  the QC  check  sample
      concentrate must be  prepared  using  stock standards  prepared independently
      from the stock standards used for calibration.

            8.5.2 Spike four pairs of  cleaned,  prepared  VOST  tubes  with  10 /zL
      of  the  QC  check sample  concentrate  and analyze  these  spiked  VOST tubes
      according to the method  beginning in Sec. 7.0.

            8.5.3 Calculate  the  average  recovery (X)  in  ng and  the standard
      deviation of the recovery (s)  in ng for  each analyte using the  results of
      the four analyses.

            8.5.4 The  average  recovery and standard deviation must fall  within
      the expected range  for determination of volatile organic compounds using
      the  VOST analytical methodology.   The expected  range  for recovery of


                                   5041  -  22                          Revision 0
                                                                September 1994

-------
      volatile  organic compounds  using  this  method  is  50-150%.    Standard
      deviation will be compound dependent, but should, in general,  range from
      15 to 30 ng.   More detailed method performance criteria must be generated
      from historical  records in the  laboratory or from interlaboratory studies
      coordinated by the  Environmental  Protection Agency.  Since the additional
      steps of  sorbent tube spiking  and desorption are  superimposed  upon the
      methodology of Method 8260,  direct transposition of Method 8260 criteria
      is questionable.  If the recovery and standard deviation for all analytes
      meet the acceptance  criteria, the system performance is acceptable and the
      analysis of field samples may begin.   If any individual standard deviation
      exceeds the precision limit  or any individual  recovery falls outside the
      range for accuracy,  then the system performance is unacceptable for that
      analyte.

            NOTE: The  large  number  of analytes listed  in Table  1  presents  a
                  substantial probability that one or  more will  fail  at least
                  one  of  the acceptance criteria when all  analytes  for this
                  method  are determined.

            8.5.5 When one or more of the  analytes tested  fails at least one of
      the acceptance criteria, the analyst  must proceed according to one of the
      alternatives below.

                  8.5.5.1      Locate  and correct the source of any problem with
            the methodology and repeat the test for  all the analytes beginning
            with Sec.  8.5.2.

                  8.5.5.2     Beginning with Sec.  8.5.2,  repeat  the test only
            for those  analytes  that  have  failed to meet  acceptance criteria.
            Repeated failure,  however,  will  confirm a general  problem  either
            with  the  measurement  system  or with  the  applicability of  the
            methodology to the particular  analyte (especially if the analyte in
            question is not listed in Table 1).   If the  problem is identified as
            originating in the measurement system,  locate and correct the source
            of the  problem  and  repeat  the test for all  compounds of interest
            beginning with Sec. 8.5.2.

      8.6   To determine  acceptable accuracy  and precision limits for surrogate
standards, the following  procedure  should  be performed.

            8.6.1 For each sample analyzed,  calculate  the percent recovery of
      each surrogate compound in the  sample.

            8.6.2 Once a minimum of thirty  samples has been analyzed, calculate
      the average percent  recovery (p) and  the standard  deviation of the percent
      recovery (s) for each of the  surrogate compounds.

            8.6.3 Calculate  the  upper  and  lower  control limits   for  method
      performance for  each surrogate standard.  This calculation is performed as
      follows:

            Upper Control  Limit (UCL) = p  + 3s
            Lower Control  Limit (LCL) = p  - 3s


                                  5041  - 23                         Revision  0
                                                                September 1994

-------
            For reference,  the  comparable  control  limits for  recovery  of the
      surrogate compounds from water and soil  in Method 8240 are:

            4-Bromofluorobenzene   Water:  86-115%     Soil:  74-121%
            l,2-Dichloroethane-d4  Water:  76-114%      Soil:  70-121%
            Toluene-d8             Water:  88-110%      Soil:  81-117%

            The control limits for the  VOST methodology would be expected to be
      similar, but exact data are not presently available. Individual laboratory
      control limits can be established by the analysis of replicate samples.

            8.6.4 If surrogate recovery is  not within the limits established by
      the laboratory,  the  following procedures  are required:  (1)  Verify that
      there  are  no errors  in  calculations,  preparation of  surrogate  spiking
      solutions, and preparation of internal standard spiking solutions.  Also,
      verify that instrument performance criteria have been met. (2) Recalculate
      the data and/or analyze a replicate sample, if replicates are available.
      (3)  If all  instrument  performance  criteria are  met  and  recovery  of
      surrogates from spiked blank  VOST  tubes (analysis  of  a method blank)  is
      acceptable, the problem is  due to  the matrix.  Emissions samples  may be
      highly acidic and may be highly loaded with target and non target organic
      compounds.  Both of  these  conditions will  affect  the  ability to recover
      surrogate compounds which are spiked  on the field samples.  The surrogate
      compound  recovery  is thus  a  valuable  indicator  of the  interactions  of
      sampled compounds  with  the matrix.    If  surrogates  spiked  immediately
      before  analysis  cannot  be  observed  with   acceptable  recovery,  the
      implications for target organic  analytes  which have been sampled  in the
      field must be assessed very carefully.  If chemical or other interactions
      are occurring on the exposed tubes, the failure to observe an analyte may
      not necessarily imply that the Destruction  and Removal  Efficiency for that
      analyte is high.

      8.7   It  is  recommended  that the laboratory  adopt  additional  quality
assurance practices for use with this method.   The  specific  practices that are
most productive depend upon the  needs  of the  laboratory  and the  nature  of the
samples analyzed.  Field duplicates may be analyzed to  assess the precision of
the environmental measurements.   When doubt exists  over the  identification of a
peak on the chromatogram, confirmatory techniques such as gas  chromatography with
a dissimilar column or a different ionization mode using a mass spectrometer may
be used,  if replicate  samples showing the same  compound are available.  Whenever
possible,  the  laboratory   should  analyze  standard reference  materials  and
participate in relevant performance evaluation studies.


9.0  METHOD PERFORMANCE

      9.1   The method detection limit  (MDL) is defined  in Chapter One.  The MDL
concentrations listed in Table 2 were obtained using cleaned blanked VOST tubes
and reagent water.  Similar results have  been achieved with  field samples.  The
MDL actually  achieved  in  a given  analysis  will  vary depending upon instrument
sensitivity and the effects of the matrix.  Preliminary spiking studies indicate
that under these conditions, the method detection limit for  spiked compounds in
extremely complex matrices may be larger by a factor of 500-1000.


                                   5041 - 24                         Revision 0
                                                                September 1994

-------
10.0  REFERENCES

1.    Protocol  for Collection  and  Analysis  of  Volatile  POHCs Using  VOST.
      EPA/600/8-84-007, March, 1984.

2.    Validation  of  the  Volatile  Organic  Sampling  Train  (VOST)  Protocol.
      Volumes I and II.  EPA/600/4-86-014A, January, 1986.

3.    U. S. EPA 40  CFR  Part  136,  "Guidelines  Establishing Test Procedures for
      Analysis of Pollutants  Under the Clean Water Act,  Method 624," October 26,
      1984.

4.    Bellar, T.  A., and J. J. Lichtenberg, J.  Amer.  Water  Works Assoc., 66(12),
      739-744, 1974.

5.    Bellar, T.  A., and J.  J. Lichtenberg, "Semi-Automated Headspace Analysis
      of Drinking Waters and  Industrial  Waters  for  Purgeable  Volatile Organic
      Compounds," in Van Hall, ed.,  Measurement of Organic Pollutants in Water
      and Wastewater,  ASTM STP 686,  pp 108-129, 1979.
                                  5041 - 25                         Revision 0
                                                                September 1994

-------
                                   TABLE 1.
       RETENTION TIMES AND CHARACTERISTIC IONS FOR VOLATILE COMPOUNDS
                     WHICH CAN BE ANALYZED BY METHOD 5041
Retention
Compound Time (min)
Acetone
Acrylonitrile
Benzene
Bromochl oromethane
Bromodichloromethane
4-Bromof 1 uorobenzene
Bromoform
Bromome thane
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chi orodi bromomethane
Chloroethane
Chloroform
Chi oromethane
Di bromomethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-l,2-Dichloroethene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,4-Difl uorobenzene
Ethyl benzene
lodomethane
Methylene chloride
Styrene
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,1,1 -Tr i chl oroethane
1 , 1 , 2 -Tri chl oroethane
Trichloroethene
Trichlorofluoromethane
1 , 2 , 3-Tr i chl oropropane
Vinyl chloride
Xylenes*
7.1
8.6
13.3
12.0
16.0
23.4
22.5
4.1
7.1
12.6
20.5
19.3
4.2
12.2
3.0
15.4
10.0
13.3
6.4
8.6
15.2
17.0
18.2
14.2
21.1
7.0
8.1
22.3
24.0
18.6
17.4
12.4
18.4
14.5
5.1
24.0
3.2
22.2
Primary Ion
Mass
43
53
78
128
83
95
173
94
76
117
112
129
64
83
50
93
63
62
96
96
63
75
75
114
106
142
84
104
83
164
92
97
97
130
101
75
62
106
Secondary Ion(s)
Mass(es)
58
52, 51
52, 77
49, 130, 51
85, 129
174, 176
171, 175, 252
96, 79
78
119, 121
114, 77
208, 206
66, 49
85, 47
52, 49
174, 95
65, 83
64, 98
61, 98
61, 98
62, 41
77, 39
77, 39
63, 88
91
127, 141
49, 51, 86
78, 103
85, 131, 133
129, 131, 166
91, 65
99, 117
83, 85, 99
95, 97, 132
103, 66
110, 77, 61
64, 61
91
The retention time  given  is  for m- and p-xylene, which  coelute  on  the wide-bore
column.  o-Xylene elutes approximately 50 seconds later.
                                   5041  -  26
    Revision 0
September 1994

-------
                                      TABLE 2.
               PRELIMINARY METHOD DETECTION LIMITS AND BOILING POINTS
                   FOR VOLATILE ORGANICS ANALYZED BY METHOD 5041*
Compound
Chloromethane
Bromomethane
Vinyl chloride
Chl oroethane
Methyl ene chloride
Acetone
Carbon disulfide
1,1-Dichloroethene
1,1-Dichloroethane
trans-l,2-Dichloroethene
Chloroform
1,2-Dichloroethane
1,1,1 -Tri chl oroethane
Carbon tetrachloride
Bromodi chloromethane
1,1, 2, 2 -Tetrachl oroethane**
1 ,2-Dichloropropane
trans - 1 , 3 -Di chl oropropene
Trichloroethene
Di bromochl oromethane
1,1,2-Trichloroethane
Benzene
cis-1, 3 -Di chl oropropene
Bromoform**
Tetrachl oroethene
Toluene
Chlorobenzene^
Ethylbenzene"*
Styrene**
Tri chl orof 1 uoromethane
lodomethane
Acrylonitrile
Dibromomethane
1 ,2,3-Trichloropropane**
total Xylenes**
CAS Number
74-87-3
74-83-9
75-01-4
75-00-3
75-09-2
67-64-1
75-15-0
75-35-4
75-35-3
156-60-5
67-66-3
107-06-2
71-55-6
56-23-5
75-27-4
79-34-5
78-87-5
10061-02-6
79-01-6
124-48-1
79-00-5
71-43-2
10061-01-5
75-25-2
127-18-4
108-88-3
108-90-7
100-41-4
100-42-5
75-69-4
74-88-4
107-13-1
74-95-3
96-18-4

Detection
Limit, ng
58
26
14
21
9
35
11
14
12
11
11
13
8
8
11
23
12
17
11
21
26
26
27
26
11
15
15
21
46
17
9
13
14
37
22
Boiling
Point, °C
-24
4
-13
13
40
56
46
32
57
48
62
83
74
77
88
146
95
112
87
122
114
80
112
150
121
111
132
136
145
24
43
78
97
157
138-144
*  The method  detection limit  (MDL)  is defined as  the minimum  concentration  of a
   substance that can be measured  and  reported  with  99% confidence that the analyte
   concentration is greater than zero and is determined from analysis of a sample in
   a given  matrix  containing  the  analyte.   The detection limits  cited  above  were
   determined according to Title 40 CFR,  Part 136, Appendix B, using standards spiked
   onto clean VOST tubes.  Since clean  VOST tubes were used,  the values cited above
   represent the best that the methodology can achieve.  The presence of an emissions
   matrix will  affect the ability of the methodology  to  perform at  its optimum level.
** Not appropriate for quantitative sampling by Method 0030.
                                      5041  -  27
    Revision 0
September 1994

-------
                                   TABLE 3.
              KEY  ION ABUNDANCE CRITERIA  FOR 4-BROMOFLUOROBENZENE
Mass                           Ion  Abundance  Criteria
 50                            15  to  40% of mass 95
 75                            30  to  60% of mass 95
 95                            base peak,  100%  relative  abundance
 96                            5 to 9% of mass  95
173                            less than 2% of  mass  174
174                            greater than 50% of mass  95
175                            5 to 9% of mass  174
176                            greater than 95%, but  less than  101%  of mass  174
177                            5 to 9% of mass  176
                                   5041  -  28                        Revision 0
                                                                September 1994

-------
                                   TABLE 4.
           VOLATILE INTERNAL STANDARDS WITH CORRESPONDING ANALYTES
                           ASSIGNED FOR  QUANTITATION
Bromochloromethane

Acetone
Acrylonitrile
Bromomethane
Carbon disulfide
Chloroethane
Chloroform
Chloromethane
1,1-Dichloroethane
1,2-Dichloroethane
l,2-Dichloroethane-d4 (surrogate)
1,1-Dichloroethene
Trichloroethene
trans-l,2-Dichloroethene
lodomethane
Methylene chloride
Trichlorofluoromethane
Vinyl chloride
1,4-Di f1uorobenzene

Benzene
Bromodi chloromethane
Bromoform
Carbon tetrachloride
Chlorodi bromomethane
Dibromomethane
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-1,3-Di chloropropene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
                               Chlorobenzene-d5
                               4-Bromofluorobenzene  (surrogate)
                               Chlorobenzene
                               Ethyl benzene
                               Styrene
                               1,1,2,2-Tetrachloroethane
                               Tetrachloroethene
                               Toluene
                               Toluene-dg  (surrogate)
                               1,2,3-Trichloropropane
                               Xylenes
                                  5041 - 29
                    Revision 0
                September 1994

-------
           o *»•
               o»
            c CTI
            o «-i
            0)
           Q£
               (1)
              OO
 o
 c
 o
 a.
 s_
 o
 
-------
           o ~*

           U)  i-
          •i-  a>
           > -Q
           CD  E
          ca  a;
              4->
               Q.
               0)
 c


 Q.
73
 C
 (O

 O)
 01
 s-
 Q.
 i.
 O
 CO
 O)
a

 0)
           ro
           O
           \f>
 O)
 s-

-------
k.
0)
1)
Vt C
ll
•* u
ID
O.
01




E
Q)
in
"o
Q




O
=5 ®
».>
2 JE
u
* J;
a><
I*
tn
i
I


c
O



j
a
0
"o
E
0
s:
o
i
a
o
V.
TJ
C
O
0)
0»
a.
/

!/>
"5
o
a.
a

i
c
o
O
                                                                                   Ol
                                                                                C <7l
                                                                                O —i

                                                                                £>  J-
                                                                                •-  Oi
                                                                                > .0
                                                                                0)  E
                                                                                OS  0)
                                                                                   -t->
                                                                                    a.
                                                                                    QJ
  O)
 4-J

  00
                                                                     U
                                                                     O)
                                                                     >
                                                                    o
 03


 Ol

•r—
Q


 O
                                                                    O)
                                                                   ^:
                                                                    u
                                                                   CO

                                                                    0)
                                                                              ro
                                                                              o
                                                                              in

-------
       Water Fill  Line
Sintered Glass Frit
                                                                         Gas  Flow
                          Figure 4.  Sample  Purge Vessel
                                     5041 - 33
    Revision  0
September  1994

-------
           o *
              en
           c 01
           o *-«
           •r-  0)
               QJ
              4->
               Q.
               0)
              00
 Q.


co

 U

 c
 rQ

 i-
o

 0)
 re

'o
>

14-
 o

 o

4->
 re

 QJ
J=
 U
oo



LO

 o>
 s-

-------
                                 METHOD  5041
PROTOCOL  FOR  ANALYSIS  OF SORBENT  CARTRIDGES FROM VOLATILE ORGANIC
        SAMPLING TRAIN:  WIDE-BORE  CAPILLARY  COLUMN TECHNIQUE

1 Start j
1
r
7.1 Conditions for
cartridge
desorption oven,
purge-and-trap
concentrator, GC,
and MS.
1
r
7.2 Daily, tune
the GC/MS with
BFB and check
calibration curve
(see Section 7.17).
>
r
7.3 - 7.6
Assemble the
system.
^
r
7.7.1 Calibrate the
instrument system
using the internal std.
procedure. Stds. and
calibration compounds
are apiked into cleaned
VOST tubes using the
flash evaporation
technique.
1
r
7.8 Prep the
purge-and-trap
unit with 5 ml
organic-free
reagent water.
1
7.9 C
paired
tubas
gas lir
desor
r
onnect
VOST
to the
tee for
ption.







7.10 Initiate
tube desorption/
purge and
heating.
J
r
7.11 Set the GC
oven to subambient
temperature
with liquid
nitrogen.
1
r
7.12 Prep the
GC/MS system
for data
aquisition.
J
r
7.13 After the tube/
water purge time,
attach the
analytical trap to
the GC/MS for
desorption.
^
r
7.14 Wash purging
vassal with two
5 mL flushes of
organic-free
reagent water.
^
r
7.15 Recondition the
analytical trap by
making it cut at
temps up to 220 C for
11 min. Trap replacemer
may be necessary
if the analytical trap
is saturated beyond
cleanup.

i
r
7.16.1 Prep
calibration stds.
                                     as in 7.7.1. Add
                                     water to vessel
                                      and desorb.
                                                                       7.16.2
                                                                     Tabulate the
                                                                    area response
                                                                   of all compounds
                                                                     of interest.
                                                                       7.16.3
                                                                     Calculate the
                                                                    average RF for
                                                                    each compound
                                                                      of interest.
                                                                    7.16.4 Calculate
                                                                      the %RSD
                                                                     for the CCCs.
                                                                    The %RSD must
                                                                      be <30%.
                                                                     7.18 GC/MS
                                                                      analysis of
                                                                       samples.
                                                                    7.19.1 Qualitative
                                                                     analysis of data
                                                                   and ident. guidelines
                                                                     of compounds.
                                                                   7.19.2 Quantitative
                                                                   analysis of data for
                                                                    the compounds of
                                                                        interest.
                                                                    (    Stop    J
                                    5041  -  35
     Revision  0
September 1994

-------
4.2  SAMPLE PREPARATION METHODS

     4.2.2  CLEANUP

         The following methods are included in this section:

         Method 3600B:     Cleanup
         Method 3610A:     Alumina Column Cleanup
         Method 3611A:     Alumina   Column   Cleanup   and  Separation   of
                           Petroleum Wastes
         Method 3620A:     Florisil Column Cleanup
         Method 3630B:     Silica Gel Cleanup
         Method 3640A:     Gel-Permeation Cleanup
         Method 3650A:     Acid-Base Partition Cleanup
         Method 3660A:     Sulfur Cleanup
         Method 3665:      Sulfuric Acid/Permanganate Cleanup
                                 FOUR - 9                        Revision  2
                                                             September 1994

-------
                                 METHOD 3600B

                                    CLEANUP
1.0   SCOPE AND APPLICATION

      1.1   Method 3600 provides general guidance on selection of cleanup methods
that are appropriate for the target analytes of interest.  Cleanup methods are
applied to the extracts prepared by one of  the extraction methods, to eliminate
sample interferences.   The following table lists the cleanup methods and provides
a brief description of the type of cleanup.

                            SW-846 CLEANUP METHODS
 Method #    Method  Name                              Cleanup Type


 3610        Alumina  Cleanup                         Adsorption

 3611        Alumina Cleanup &  Separation             Adsorption
             for  Petroleum  Waste

 3620        Florisil   Cleanup                        Adsorption

 3630        Silica  Gel   Cleanup                      Adsorption

 3640        Gel-Permeation Cleanup                   Size-Separation

 3650        Acid-Base Partition  Cleanup              Acid-Base Partitioning

 3660        Sulfur  Cleanup                          Oxidation/Reduction

 3665        Sulfuric  Acid/Permanganate               Oxidation/Reduction
             Cleanup


      1.2   The purpose of applying a cleanup method to an extract is to remove
interferences  and  high boiling material  that  may result  in:    (1)  errors in
quantitation (data  may be  biased low  because of  analyte  adsorption  in  the
injection port  or  front of the  GC  column or biased high because of overlap with
an interference peak);  (2) false positives because  of interference peaks falling
within the analyte retention time window; (3)  false negatives caused by shifting
the  analyte  outside  the  retention  time  window;  (4)  rapid deterioration of
expensive capillary  columns; and,  (5) instrument downtime  caused by cleaning and
rebuilding of detectors and  ion sources.  Most extracts of soil and waste require
some degree of  cleanup, whereas, cleanup for water extracts may be unnecessary.
Highly  contaminated  extracts  (e.g.   sample  extracts   of oily  waste or  soil
containing oily residue)  often  require a combination  of  cleanup  methods.   For
example,  when  analyzing  for  organochlorine pesticides  and PCBs,  it may be
necessary to use  gel  permeation  chromatography  (GPC),  to eliminate  the  high
boiling  material   and  a  micro  alumina  or  Florisil  column  to  eliminate
interferences with the analyte peaks on the GC/ECD.

                                   3600B -  1                         Revision 2
                                                                September 1994

-------
      1.3   The following techniques have  been applied to extract purification:
adsorption  chromatography;  partitioning  between  immiscible  solvents;  gel
permeation chromatography; oxidation of interfering substances with acid, alkali,
or oxidizing agents.  These  techniques  may  be  used  individually or in various
combinations, depending on the extent and nature of the co-extractives.

            1.3.1       Adsorption column chromatography - Alumina (Methods 3610
      and 3611),  Florisil  (Method  3620), and silica gel  (Method 3630) are useful
      for separating analytes of  a relatively  narrow  polarity  range away from
      extraneous,  interfering  peaks  of  a  different  polarity.    These  are
      primarily  used  for cleanup  of  a  specific chemical group  of relatively
      non-polar analytes, i.e.,  organochlorine pesticides, polynuclear aromatic
      hydrocarbons  (PAHs),   nitrosamines,   etc..     Solid   phase   extraction
      cartridges have been added as an option.

            1.3.2       Acid-base  partitioning   (Method  3650)  -  Useful  for
      separating acidic or basic  organics from  neutral  organics.   It has been
      applied to analytes such as  the  chlorophenoxy herbicides and phenols.  It
      is very useful  for separating the neutral PAHs  from  the  acidic phenols
      when analyzing a site contaminated with creosote and pentachlorophenol.

            1.3.3       Gel  permeation chromatography  (GPC)  (Method 3640) - The
      most  universal  cleanup  technique  for a  broad  range of  semivolatile
      organics   and   pesticides.      It   is   capable   of  separating   high
      molecular-weight,  high boiling material from the  sample analytes.   It has
      been used  successfully  for  all  the  semivolatile  base, neutral,  and acid
      compounds  associated with  the EPA Priority Pollutant  and  the Superfund
      Target  Compound  list  prior  to  GC/MS  analysis  for   semivolatiles  and
      pesticides.  GPC may not be applicable to  elimination of  extraneous peaks
      on a chromatogram which interfere with the analytes of interest.   It is,
      however, useful  for  the removal  of high boiling  materials  which  would
      contaminate  injection  ports and  column  heads,  prolonging column  life,
      stabilizing the instrument, and reducing  column reactivity.

            1.3.4       Sulfur cleanup  (Method  3660)  -  Useful  in  eliminating
      sulfur from sample extracts, which may cause chromatographic interference
      with analytes of interest.

      1.4   Several of the methods are also  useful for fractionation of complex
mixtures of analytes.   Use the solid phase extraction cartridges in Method 3630
(Silica Gel) for separating the PCBs away from most organochlorine pesticides.
Method 3611 (Alumina)  is for  the  fractionation of aliphatic, aromatic and polar
analytes.  Method 3620 (Florisil) provides fractionation of the organochlorine
pesticides.

      1.5   Cleanup capacity is  another  factor  that must  be considered  in
choosing a  cleanup  technique.   The adsorption  methods  (3610,  3620,  and  3630)
provide the option of using standard column chromatography techniques or solid
phase extraction cartridges.   The  decision process in  selecting  between the
different options available generally depends on the amount of interferences/high
boiling material  in the  sample extract and the degree of cleanup  required  by the
determinative method.  The solid phase  extraction cartridges require  less elution
solvent and  less  time,  however,  their cleanup  capacity is drastically reduced
when comparing a 0.5 g or 1.0 g Florisil cartridge to a 20 g standard Florisil

                                   3600B -  2                        Revision 2
                                                                September 1994

-------
column.  The same factor enters into the  choice of the 70 g gel permeation column
specified in Method 3640 versus a high efficiency column.

      1.6   Table  1  indicates  the  recommended  cleanup   techniques  for  the
indicated groups of compounds.  This information  can also be used as guidance for
compounds that are not listed.  Compounds that are chemically similar to these
groups  of  compounds  should  behave similarly when  taken  through  the cleanup
procedure,however, this must be demonstrated by determining recovery of standards
taken through the method.


2.0   SUMMARY OF METHOD

      2.1   Refer to the  specific cleanup method for a summary of the procedure.


3.0   INTERFERENCES

      3.1   Analytical interferences may be caused by  contaminants in solvents,
reagents,  glassware,   and  other sample  processing  hardware.    All   of  these
materials must be routinely demonstrated to be free of interferences, under the
conditions of the analysis, by running laboratory reagent blanks.

      3.2   More extensive procedures than  those outlined in the methods may be
necessary for reagent purification.


4.0   APPARATUS AND MATERIALS

      4.1   Refer to  the  specific  cleanup  method for apparatus  and  materials
needed.
5.0   REAGENTS

      5.1   Refer to the specific cleanup method for the reagents needed.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory  material  to this  chapter,  Organic Analytes,
Section 4.1.


7.0   PROCEDURE

      7.1   Prior to  using  the cleanup  procedures, samples  normally  undergo
solvent  extraction.   Chapter  Two,  Section 2.0,  may  be used  as a  guide  for
choosing the appropriate extraction  procedure based  on the physical composition
of the waste and on the analytes of interest in the matrix (see also Method 3500
for a  general  description  of the  extraction  technique).    For  some  organic
liquids, extraction prior to cleanup may not be  necessary.
                                  3600B  - 3                         Revision 2
                                                                September 1994

-------
      7.2   Most soil/sediment and waste sample extracts will require some degree
of cleanup.  The extract is then analyzed by one of the determinative methods.
If interferences still preclude analysis for the analytes of  interest, additional
cleanup may be required.

      7.3   Many of the determinative methods specify cleanup methods that should
be  used   when   determining  particular  analytes   (e.g.   Method   8061,   gas
chromatography of phthalate esters, recommends using either  Method 3610 (Alumina
column cleanup)  or Method 3620 (Florisil column cleanup) if interferences prevent
analysis.   However,  the  experience of  the  analyst  may prove  invaluable  in
determining which cleanup  methods are  needed.   As  indicated in Section 1.0 of
this method, many matrices may  require a combination of cleanup procedures in
order to ensure proper analytical determinations.

      7.4   Guidance  for cleanup  is specified  in  each of the  methods  that follow.
The amount of extract cleanup required  prior to  the final determination depends
on the concentration  of  interferences in the sample, the selectivity of both the
extraction procedure and  the determinative method and  the required  detection
limit.

      7.5   Following cleanup,  the  sample is concentrated to whatever volume is
required  in  the determinative method.    Analysis follows as specified  in the
determinative procedure.


8.0   QUALITY CONTROL

      8.1   Refer to  Chapter One for specific quality control  procedures.

      8.2   The analyst must  demonstrate that  the   compounds  of interest are
being quantitatively recovered by the cleanup technique  before the cleanup is
applied  to actual samples.   For  sample  extracts that are  cleaned  up,  the
associated  quality  control samples   (e.g.  spikes,   blanks,   replicates,  and
duplicates) must also be processed through the same cleanup procedure.

      8.3   The  analysis   using  each  determinative  method  (GC, GC/MS,  HPLC)
specifies  instrument calibration  procedures  using  stock   standards.   It  is
recommended  that  cleanup   also be  performed  on   a  series of the same  type  of
standards  to  validate  chromatographic elution  patterns for the compounds  of
interest and to verify the  absence of  interferences from reagents.


9.0   METHOD PERFORMANCE

      9.1   Refer to  the specific cleanup method for performance data.


10.0  REFERENCES

      10.1  Refer to the specific cleanup method.
                                   3600B -  4                         Revision 2
                                                                September 1994

-------
                                   TABLE 1.
       RECOMMENDED CLEANUP TECHNIQUES FOR INDICATED GROUPS OF COMPOUNDS
Analyte Group
Determinative8
   Method
   Cleanup
Method Options
Phenols
Phthalate esters
Nitrosamines
Organochlorine pesticides & PCBs
PCBs
Nitroaromatics and cyclic ketones
Polynuclear aromatic hydrocarbons
Chlorinated hydrocarbons
Organophosphorus pesticides
Chlorinated herbicides
Semivolatile organics
Petroleum waste
PCDDs and PCDFs by LR/MS
PCDDs and PCDFs by HR/MS
N-methyl carbamate pesticides
8040
8060/8061
8070
8080/8081
8080/8081
8090
8100/8310
8120/8121
8140/8141
8150/8151
8250/8270
8250/8270
8280
8290
8318
3630b, 3640, 3650, 8040°
3610, 3620, 3640
3610, 3620, 3640
3620, 3640, 3660
3665
3620, 3640
3611, 3630, 3640
3620, 3640
3620
8150d, 8151d, 3620
3640, 3650, 3660
3611, 3650
8280
8290
8318
    The  GC/MS Methods, 8250 and 8270,  are  also appropriate determinative methods
    for  all  analyte  groups,  unless  lower  detection  limits  are  required.

b   Cleanup  applicable  to derivatized phenols.

0   Method 8040 includes a derivatization  technique  followed by GC/ECD analysis,
    if interferences are  encountered  using  GC/FID.

d   Methods  8150  and 8151 incorporate an  acid-base  cleanup step as an integral
    part of  the methods.
                                  3600B - 5
                                Revision  2
                            September 1994

-------
                    METHOD  3600B
                       CLEANUP
       Start
        \
       7.1
    Do solvent
     extraction
        1
        7.2
  Analyze analyte
 by a determinative
method from Sec. 4.3
      7.2 Are
     analytes
   undeterminable
      due to
    nterferenoe?
        7.3
 Use cleanup method
  specified for the
determinative method
                                         7.5
                                  Concentrate sample
                                   to required volume
                         3600B  -  6
                                 Revision 2
                            September  1994

-------
                                 METHOD 3610A

                            ALUMINA COLUMN CLEANUP
 1.0   SCOPE AND APPLICATION

      1.1   Scope:   Alumina is a highly  porous  and  granular form of  aluminum
 oxide.  It is available in three pH ranges (basic,  neutral,  and  acidic)  for use
 in  column  chromatography.   It is used to  separate  analytes from interfering
 compounds of a different  chemical polarity.

      1.2   General Applications  (Gordon  and Ford):

            1.2.1 Basic (B) pH (9-10): USES: Basic  and neutral compounds stable
      to alkali, alcohols, hydrocarbons, steroids,  alkaloids, natural pigments.
      DISADVANTAGES:  Can  cause polymerization,  condensation,  and dehydration
      reactions; cannot use acetone or  ethyl acetate  as eluants.

            1.2.2 Neutral  (N):  USES:   Aldehydes,  ketones,   quinones,   esters,
      lactones, glycoside.   DISADVANTAGES: Considerably  less  active than the
      basic form.

            1.2.3 Acidic  (A)  pH  (4-5):  USES:  Acidic  pigments  (natural  and
      synthetic), strong  acids (that  otherwise  chemisorb to neutral  and basic
      alumina).

            1.2.4 Activity  grades:  Acidic, basic, or neutral   alumina  can  be
      prepared in various activity grades  (I to V), according to the Brockmann
      scale, by addition  of water to Grade  1 (prepared by heating at 400-450°C
      until no more water is  lost).   The  Brockmann scale (Gordon and Ford,  p.
      374) is reproduced  below:

      Water added (wt. %):      03          6      10         15
      Activity grade:            I          II         III     IV         V
      RF (p-aminoazobenzene):   0.0       0.13       0.25    0.45       0.55

      1.3   Specific applications:  This method includes guidance for cleanup of
 sample extracts containing phthalate esters and nitrosamines.  For alumina column
 cleanup of petroleum wastes, see Method 3611, Alumina Column Cleanup of Petroleum
 Wastes.
2.0   SUMMARY OF METHOD

      2.1   The column is packed with the required amount of adsorbent, topped
with a water adsorbent,  and then loaded with the sample to be analyzed.  Elution
of the analytes is effected with a  suitable solvent(s), leaving the interfering
compounds on the column.  The eluate is then concentrated (if necessary).
                                   3610A  -  1                         Revision 1
                                                                     July 1992

-------
3.0   INTERFERENCES

      3.1   A reagent blank should be performed for  the  compounds of interest
prior to the use of this method.  The level of interferences must be below the
method detection limit before this method is performed on actual  samples.

      3.2   More extensive procedures  than  those outlined in this method may be
necessary for reagent purification.


4.0   APPARATUS AND MATERIALS

      4.1   Chromatography column:  300  mm  x 10 mm ID, with Pyrex glass wool at
bottom and a Teflon stopcock.

      NOTE:       Fritted  glass  discs  are  difficult  to decontaminate  after
                  highly  contaminated  extracts  have  been  passed  through.
                  Columns without  frits may be purchased.   Use  a small  pad of
                  Pyrex glass wool to retain the adsorbent.   Prewash the glass
                  wool pad with  50 mL of acetone  followed by 50 mL of elution
                  solvent prior  to packing  the column with adsorbent.

      4.2   Beakers:  500 mL.

      4.3   Reagent bottle:   500 mL.

      4.4   Muffle furnace.

      4.5   Kuderna-Danish (K-D) apparatus:

            4.5.1 Concentrator tube:  10 mL, graduated  (Kontes K-570050-1025 or
      equivalent).   A ground-glass stopper  is used  to  prevent  evaporation of
      extracts.

            4.5.2 Evaporation   flask:     500   mL   (Kontes   K-570001-500  or
      equivalent).    Attach  to  concentrator tube  with  springs,  clamps,  or
      equivalent.

            4.5.3 Snyder  column:  Three  ball  macro  (Kontes  K-503000-0121 or
      equivalent).

            4.5.4 Snyder  column:   Two  ball micro  (Kontes  K-569001-0219 or
      equivalent).

            4.5.5 Springs  -   1/2 inch (Kontes K-662750 or equivalent).

      4.6   Boiling chips: Solvent extracted, approximately  10/40 mesh (silicon
carbide or equivalent).

      4.7   Water  bath:    Heated,  with  concentric  ring   cover,  capable of
temperature  control  (±5°C).  The bath should be used in a hood.

      4.8   Vials:  Glass, 2 mL capacity, with Teflon lined  screw caps or crimp
tops.

                                   3610A -  2                         Revision  1
                                                                     July 1992

-------
      4.9   Erlenmeyer flasks:  50 and 250 mL


5.0   REAGENTS

      5.1   Organic-free reagent water -  All  references to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.2   Sodium sulfate:     Sodium sulfate  (granular,  anhydrous),  Na2S04.
Purify by heating at 400°C for 4 hours  in  a shallow tray, or by precleaning the
sodium sulfate with methylene chloride.  If the sodium sulfate is precleaned with
methylene chloride,  a method blank must be analyzed, demonstrating that there is
no interference from the sodium sulfate.

      5.3   Eluting  solvents:

            5.3.1 Diethyl  Ether,  C2H5OC2H5.   Pesticide quality or equivalent.
      Must  be free   of  peroxides as  indicated  by test  strips  (EM  Quant,  or
      equivalent).   Procedures  for  removal of  peroxides  are provided with the
      test strips.  After cleanup, 20 ml of ethyl alcohol preservative must be
      added to each liter of ether.

            5.3.2 Methanol, CH3OH - Pesticide quality or equivalent.

            5.3.3 Pentane, CH3(CH2)3CH3 - Pesticide quality or equivalent.

            5.3.4 Hexane, C6H14  -  Pesticide quality or equivalent.

            5.3.5 Methylene chloride, CH2C12  - Pesticide  quality  or equivalent.

      5.4   Alumina;

            5.4.1 For cleanup of phthalate extracts:  Alumina-Neutral, activity
      Super I, W200 series (ICN  Life Sciences  Group, No. 404583, or equivalent).
      To prepare for use, place 100 g of alumina into a 500 ml beaker and heat
      for approximately  16 hr at 400°C.   After heating, transfer  to a 500 ml
      reagent bottle.  Tightly  seal  and  cool to room temperature.  When cool,
      add 3 ml of organic-free reagent  water.   Mix thoroughly  by shaking or
      rolling for 10 min  and  let  it  stand for  at  least  2 hr.   Keep the bottle
      sealed tightly.

            5.4.2 For cleanup of nitrosamine  extracts:  Alumina-Basic, activity
      Super I, W200 series (ICN  Life Sciences  Group, No. 404571, or equivalent).
      To prepare for use, place 100  g  of alumina  into  a  500 mL  reagent bottle
      and add 2 mL of organic-free reagent water.  Mix the alumina preparation
      thoroughly by shaking or rolling for 10 min  and let it stand  for  at least
      2 hr.   The preparation  should  be  homogeneous before use.  Keep  the bottle
      sealed tightly to ensure proper activity.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Section 4.1.

                                  3610A  - 3                         Revision 1
                                                                     July 1992

-------
7.0   PROCEDURE

      7.1   Phthalate esters:

            7.1.1 Reduce the  sample  extract  volume to 2 ml prior  to cleanup.
      The extract solvent must be hexane.

            7.1.2 Place  approximately  10 g  of  alumina  into  a  10  mm  ID
      chromatographic column.   Tap the  column to  settle the alumina and add 1-2
      cm of anhydrous sodium sulfate to the top.

            7.1.3 Pre-elute the column with 40 ml of hexane.   The rate for all
      elutions should be about 2 mL/min.   Discard the eluate and, just prior to
      exposure of the sodium sulfate layer to the air,  quantitatively transfer
      the 2 ml sample extract onto the column  using  an additional 2 ml of hexane
      to complete the transfer.   Just  prior  to exposure  of the sodium sulfate
      layer to the  air,  add 35 ml of  hexane and continue the  elution  of the
      column.   Discard this hexane eluate.

            7.1.4 Next,  elute  the column  with 140 ml of  20% ethyl  ether in
      hexane  (v/v) into  a  500  mL  K-D flask equipped with a 10 mL concentrator
      tube.    Concentrate   the collected  fraction using  the  Kuderna-Danish
      technique.   No  solvent exchange  is  necessary.  Adjust the volume  of the
      cleaned  up  extract to whatever  volume  is  required  (10.0 ml  for  Method
      8060) and analyze.  Compounds that elute  in this fraction are as follows:

            Bis(2-ethylhexyl)  phthalate
            Butyl benzyl phthalate
            Di-n-butyl phthalate
            Diethyl  phthalate
            Dimethyl  phthalate
            Di-n-octyl phthalate.

      7.2   Nitrosamines:

            7.2.1 Reduce the sample  extract to 2 ml prior  to cleanup.

            7.2.2 Diphenylamine,  if present in  the original  sample extract, must
      be  separated  from the  nitrosamines if  N-nitrosodiphenylamine  is  to be
      determined by this method.

            7.2.3 Place  approximately  12 g of the alumina preparation  into  a 10
      mm  ID chromatographic column.  Tap  the column to  settle the alumina and
      add  1-2  cm of anhydrous  sodium sulfate to  the top.

            7.2.4 Pre-elute the  column   with  10  ml  of  ethyl  ether/pentane
      (3:7)(v/v).  Discard  the eluate  (about 2 ml) and,  just prior to exposure
      of  the  sodium  sulfate layer to the  air,  quantitatively transfer the 2 ml
      sample  extract onto the  column  using  an additional  2 ml  of pentane to
      complete the transfer.

            7.2.5 Just prior to exposure  of the sodium sulfate layer to the  air,
      add  70  ml  of ethyl ether/pentane (3:7)(v/v).  Discard the first 10 ml of
      eluate.    Collect  the remainder of the eluate  in  a  500 mL  K-D flask

                                   3610A  - 4                         Revision 1
                                                                     July  1992

-------
      equipped with  a 10 ml  concentrator tube.   This  fraction contains   N-
      nitroso-di-n-propylamine.

            7.2.6 Next,  elute the  column with 60  ml of  ethyl  ether/pentane
      (l:l)(v/v), collecting  the eluate  in  a  second 500  mL K-D flask equipped
      with a 10 ml concentrator tube.  Add 15 mL of methanol to the K-D flask.
      This fraction will  contain  N-nitrosodimethylamine, most of the N-nitroso-
      di-n-propylamine, and any diphenylamine that is present.

            7.2.7 Concentrate both fractions using  the Kuderna-Danish Technique
      (if  necessary),  using  pentane  to  prewet the  Snyder column.   When  the
      apparatus is cool, remove  the Snyder  column  and rinse the flask and its
      lower joint into the concentrator tube with 1-2 ml of pentane.  Adjust the
      final volume  to whatever  is  required in the  appropriate determinative
      method (Section 4.3 of  this chapter).   Analyze the fractions.


8.0   QUALITY CONTROL

      8.1   Refer to  Chapter  One for specific quality control  procedures  and
Method 3600 for cleanup procedures.

      8.2   The analyst should demonstrate  that the compounds  of interest  are
being quantitatively recovered before applying this method to actual samples.

      8.3   For sample  extracts  that are cleaned  up  using this  method,  the
associated quality control  samples must also be processed through this cleanup
method.
9.0   METHOD PERFORMANCE

      9.1   Performance data are not available.


10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act;  Final  Rule and Interim
      Final Rule and Proposed Rule," October 26,  1984.
                                  3610A  - 5                         Revision 1
                                                                     July 1992

-------
                        METHOD 3610A
                 ALUMINA COLUMN CLEANUP
   711  Reduce
     volume of
      sample
     extract
   7.1  2  Put
  alumina in
  column,  add
   anhydrous
sodium  sulfate
       7  1  3
     Preclute
    column  with
      hexane
  713  Transfer
  sample ex tract
    to column,
   elute column
   with  hexane.
714  Elute column
    with  ethyl
   ether/hexane
 Collect  eluate in
      flask
                                                   7 2 1 Reduce
                                                    volume of
                                                     sample
                                                    extract
                                                      7 2 3 Put
                                                     alumina in
                                                     column, add
                                                      anhydrous
                                                    sodium sulfate
                                                  724 Preelute
                                                 column with ethyl
                                                  ether/pentane
                                                  Transfer sample
                                                extract to column,
                                                   add pentane
                                                725 Elute column
                                                    with ethyl
                                                  ether/pentane
                                                 Collect eluate in
                                                     flask
                                                7  2  6 Elute column
                                                    with ethyl
                                                  ether/pentane
                                                Collect eluate in
                                                second flask, add
                                                    methanol
     714
  Concentrate
   collected
   fraction,
adjust  volume
                             Analyze by
                             appropriate
                            determinative
                               method
                                                       7 2 ?
                                                    Concentrate
                                                  both  fractions,
                                                  ad jus t volume
                            3610A -  6
                                                                       Revision  1
                                                                        July 1992

-------
                                 METHOD 3611A

           ALUMINA COLUMN CLEANUP AND  SEPARATION  OF  PETROLEUM  WASTES
1.0   SCOPE AND APPLICATION

      1.1   Method 3611 was formerly Method 3570 in the Second Edition of this
manual.

      1.2   Specific application:  This  method  includes guidance for separation
of petroleum wastes into aliphatic, aromatic, and polar fractions.


2.0  SUMMARY OF METHOD

      2.1  The column  is packed  with  the  required  amount of adsorbent,  topped
with a water adsorbent, and then  loaded with the sample to be analyzed.  Elution
of the analytes is effected with  a  suitable solvent(s), leaving the interfering
compounds on the column.  The eluate is then concentrated (if necessary).


3.0  INTERFERENCES

      3.1   A reagent  blank should  be performed  for  the  compounds of interest
prior to the use of this method.   The level of interferences must be below the
method detection limit before this method is performed on actual  samples.

      3.2   More extensive procedures  than those  outlined in this method may be
necessary for reagent purification.

      3.3   Caution must be taken to prevent overloading of the chromatographic
column.  As the column loading for any of these types of wastes approaches
0.300  g of  extractable  organics, separation  recoveries  will   suffer.    If
overloading is suspected, an aliquot of the base-neutral extract prior to cleanup
may be weighed and then evaporated to dryness.   A gravimetric determination on
the aliquot will  indicate the weight of extractable organics in the sample.

      3.4   Mixtures of petroleum wastes containing predominantly polar solvents,
i.e., chlorinated solvents or  oxygenated solvents, are not appropriate for this
method.


4.0  APPARATUS AND MATERIALS

      4.1  Chromatography column:  300 mm x 10 mm ID, with Pyrex glass wool at
bottom and a Teflon stopcock.

      NOTE: Fritted glass discs  are difficult  to  decontaminate  after  highly
            contaminated extracts  have  been  passed through.   Columns without
            frits may  be purchased.   Use a  small  pad of Pyrex glass wool  to
            retain the  adsorbent.   Prewash the  glass  wool  pad with  50  mL of
            acetone followed by  50 mL of  elution solvent prior to packing the
            column with adsorbent.

                                   3611A -  1                         Revision 1
                                                                     July 1992

-------
      4.2   Beakers:   500 ml.

      4.3   Reagent bottle:   500 ml.

      4.4   Muffle furnace.

      4.5   Kuderna-Danish (K-D) apparatus:

            4.5.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
      equivalent).  A ground-glass stopper  is used to prevent  evaporation of
      extracts.

            4.5.2 Evaporation   flask   -   500  ml   (Kontes  K-570001-500   or
      equivalent).   Attach  to  concentrator  tube  with   springs,  clamps,  or
      equivalent.

            4.5.3 Snyder column  -  Three ball  macro (Kontes K-503000-0121 or
      equivalent).

            4.5.4 Snyder  column - Two ball  micro  (Kontes  K-569001-0219 or
      equivalent).

            4.5.5 Springs - 1/2 inch  (Kontes K-662750 or  equivalent).

      4.6   Boiling chips:  Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.7   Water  bath:    Heated  with  concentric  ring  cover,  capable  of
temperature control (±5°C).   The bath  should be used in a hood.

      4.8   Erlenmeyer flasks:  50 and 250 mL.


5.0  REAGENTS

      5.1   Sodium sulfate:  (granular,  anhydrous), Na2S04.  Purify by heating at
400°C for 4 hours in a shallow tray,  or by precleaning the sodium sulfate with
methylene chloride. If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed, demonstrating that there is no interference  from
the sodium sulfate.

      5.2   Eluting solvents:

            5.2.1 Methanol, CH3OH - Pesticide  quality or  equivalent.

            5.2.2 Hexane, C6HU  - Pesticide  quality or equivalent.

            5.2.3 Methylene chloride,  CH2C12 - Pesticide quality  or equivalent.

      5.3   Alumina:  Neutral  80-325  MCB  chromatographic grade or equivalent.
Dry alumina overnight at 130°C prior  to use.
                                   3611A -  2                        Revision 1
                                                                     July 1992

-------
6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory material  to this chapter,  Organic Analytes,
Section 4.1.
7.0  PROCEDURE

      7.1   It is suggested that  Method  3650,  Acid-Base Partition Cleanup, be
performed on the sample extract prior to alumina cleanup.

      7.2   Place approximately 10 g of alumina into a chromatographic column,
tap to settle the alumina,  and add 1  cm of anhydrous sodium sulfate to the top.

      7.3   Pre-elute the column with 50  mL  of  hexane.  Discard the eluate and,
just prior to exposure of  the sodium sulfate layer to the air, quantitatively
transfer the  1 mL  sample extract onto the column  using  an  additional  1  mL of
hexane  to  complete  the  transfer.   To  avoid  overloading  the column,  it is
suggested that no more  than  0.300 g of  extractable organics  be  placed on the
column (see Section 3.3).

      7.4   Just prior to exposure of the sodium sulfate to the air, elute the
column with a total  of 15 mL of hexane.  If the extract is in 1  mL of hexane, and
if 1 mL of  hexane was used  as  a rinse, then 13 mL of additional  hexane should be
used.    Collect  the  effluent  in  a 50 mL  flask  and  label  this  fraction
"base/neutral aliphatics."  Adjust the flow rate to 2 mL/min.

      7.5   Elute the column with 100 mL of methylene chloride and collect the
effluent in a 250 mL flask.   Label this fraction "base/neutral aromatics."

      7.6   Elute the column with  100 mL of methanol and  collect the effluent in
a 250 mL flask.   Label  this fraction "base/neutral polars."

      7.7   Concentrate the extracts  (if necessary)  by the standard  K-D technique
to the volume  (1-10 mL)  required in the appropriate  determinative method (Chapter
Four).  Analyze the fractions containing the analytes of interest.


8.0  QUALITY CONTROL

      8.1   Refer to Chapter  One for specific quality  control  procedures and
Method 3600 for cleanup procedures.

      8.2   The analyst  should demonstrate  that the  compounds of interest are
being quantitatively recovered before applying  this method to actual samples.

      8.3   For  sample  extracts  that  are  cleaned up  using  this  method,  the
associated quality control samples must also be processed through this cleanup
method.
                                   3611A  -  3                         Revision 1
                                                                     July 1992

-------
9.0   METHOD PERFORMANCE

      9.1   The precision and accuracy of the method will depend upon the overall
performance of the sample preparation and analysis.

      9.2   Rag oil  is an emulsion  consisting  of crude oil,  water,  and soil
particles.  It has a density greater than crude oil and less than water.  This
material forms a  layer between the crude oil  and water when  the crude oil  is
allowed to gravity separate at the refinery.  A rag oil sample was analyzed by
a number of laboratories according to the procedure outlined in this method.  The
results of these  analyses by GC/MS for  selected components in the rag oil are
presented in Table 1.  Reconstructed ion chromatograms from the GC/MS analyses
are included as Figures 1 and 2.


10.0  REFERENCES

1.    U.S. EPA 40  CFR Part 136, "Guidelines Establishing Test Procedures  for the
      Analysis of Pollutants Under the Clean Water Act; Final  Rule and  Interim
      Final Rule and Proposed Rule," October 26, 1984.
                                   3611A - 4                         Revision  1
                                                                      July 1992

-------
                                   Table 1.
            RESULTS OF ANALYSIS FOR SELECTED COMPONENTS IN RAG OIL
                                        Mean              Standard
Analyte                             Cone. (mg/kg)a       Deviation      %RSD


Naphthalene                                216              42           19
Fluorene                                   140              66           47
Phenanthrene                               614             296           18
2-Methylnaphthalene                        673             120           18
Dibenzothiophene                          1084             286           26
Methylphenanthrene                        2908            2014           69
Methyldibenzothiophene                    2200            1017           46


                                    Average Surrogate Recovery

Nitrobenzene-d5                           58.6            11
Terphenyl-du                              83.0             2.6
Phenol-d6                                 80.5            27.6
Naphthalene-d8                            64.5             5.0
a  Based on five determinations from three laboratories.

b  Percent Relative Standard Deviation.
                                  3611A  - 5                         Revision 1
                                                                     July 1992

-------
Figure 1
 3611A - 6
Revision 1
 July 1992

-------
                               Figure 2
ts
XI

II
      u
      fO

      V
                               3611A - 7
Revision 1
 July 1992

-------
                           METHOD 3611A
ALUMINA  COLUMN CLEANUP AND SEPARATION OF  PETROLEUM  WASTES
                       Start
                    7 1  Cleanup
                   using Method
                       3650
                   7 2 Add alumina
                        to
                   chroma tographic
                      column
                      7  2 Add
                     anhydrous
                  sodium sulfate
                     to  top of
                      column
                   7 3 Preelute
                    column with
                      hexane
                        7 3
                  Quantitatively
                  add extract to
                      column





















7 4 Elute
"base-neutral
al iphatics "
fraction wi th
hexi

me

7 5 Elute
"base-neutral
ar omatics "
fraction with
CH2C12


7 6 Elute
"base-neutral
polars"
fraction wi th
met Hanoi

7

7
Concent ra te
ex tracts


s^ "^
/ Analyze using
/ appropriate
                                           method
                            3611A  -  8
Revision  1
 July 1992

-------
                                 METHOD 3620A

                            FLORISIL  COLUMN  CLEANUP
1.0   SCOPE AND APPLICATION

      1.1   Florisil, a registered trade name of the Floridin Co.,  is a magnesium
silicate with acidic properties.  It is used for general  column chromatography
as a cleanup procedure prior to sample analysis  by gas chromatography.

      1.2   General  applications:   Cleanup  of  pesticide  residues and  other
chlorinated hydrocarbons; the separation of nitrogen compounds from hydrocarbons;
the  separation  of  aromatic  compounds  from  aliphatic-aromatic mixtures;  and
similar  applications   for   use  with  fats,  oils,   and   waxes  (Floridin).
Additionally, Florisil  is considered good for separations with steroids, esters,
ketones, glycerides, alkaloids, and some carbohydrates (Gordon and Ford).

      1.3   Specific applications:  This method includes guidance  for cleanup of
sample  extracts  containing  the following analyte  groups: phthalate esters;
nitrosamines; organochlorine pesticides; nitroaromatics; haloethers; chlorinated
hydrocarbons; and organophosphorus pesticides.


2.0   SUMMARY OF METHOD

      2.1   The column is packed with the required adsorbent, topped with a water
adsorbent,  and then loaded  with the  sample to be analyzed.  Elution is effected
with a suitable  solvent(s) leaving the interfering compounds on  the column.  The
eluate is then concentrated (if necessary).


3.0   INTERFERENCES

      3.1   A reagent  blank  should  be  performed  for the  compounds of interest
prior to the use of this method.  The level  of interferences must be below the
method detection limit before this method is performed on actual  samples.

      3.2   More extensive procedures  than those outlined in this method may be
necessary for reagent purification.


4.0   APPARATUS AND MATERIALS

      4.1   Beaker - 500 mL.

      4.2   Chromatographic column  - 300 mm  long  x 10 mm  ID or  400  mm long x
20 mm ID, as  specified in  Section  7.0; with  Pyrex glass wool  at bottom and a
Teflon stopcock.

      NOTE: Fritted  glass  discs are difficult  to decontaminate  after highly
            contaminated extracts  have been passed through.   Columns without
            frits may  be  purchased.   Use a  small  pad of  Pyrex  glass wool  to
            retain the  adsorbent.   Prewash  the glass  wool pad with  50  mL  of

                                   3620A - 1                         Revision 1
                                                                     July 1992

-------
            acetone followed by 50 ml of elution solvent prior  to  packing the
            column with adsorbent.

      4.3   Kuderna-Danish (K-D)  apparatus.

            4.3.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
      equivalent).   Ground-glass stopper  is  used to  prevent  evaporation  of
      extracts.

            4.3.2 Evaporation  flask  -   500   ml  (Kontes   K-570001-0500  or
      equivalent).   Attach  to  concentrator  tube  with springs,  clamps,  or
      equivalent.

            4.3.3 Snyder column  -  Three ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.3.4 Snyder  column  -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.3.5 Springs -  1/2 inch (Kontes  K-662750 or equivalent).

      4.4   Muffle furnace.

      4.5   Reagent bottle - 500 ml.

      4.6   Water  bath  -  Heated,  with  concentric   ring  cover,  capable  of
temperature control (±5°C).  The  bath should be used  in  a hood.

      4.7   Boiling chips - Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.8   Erlenmeyer flasks - 50 and 250 ml.

      4.9   Top-loading balance - 0.01 g.


5.0   REAGENTS

      5.1   Organic-free reagent water - All references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.2   Florisil  -  Pesticide residue  (PR)  grade (60/100 mesh);  purchase
activated  at  1250°F  (677°C),  stored  in  glass  containers  with  ground-glass
stoppers or foil-lined screw caps.

            5.2.1 Deactivation of  Florisil  -  for cleanup of phthalate esters.
      To prepare for use, place 100 g of Florisil  into a 500  ml  beaker and heat
      for  approximately  16 hr at  40°C.   After heating, transfer  to  a  500 ml
      reagent bottle.  Tightly seal  and cool to room temperature. When cool add
      3 ml of organic-free reagent water.   Mix thoroughly by  shaking or rolling
      for  10  min and  let  stand for  at  least 2 hr.    Keep  the bottle  sealed
      tightly.

            5.2.2 Activation  of  Florisil  -  for  cleanup  of nitrosamines,

                                   3620A  - 2                        Revision  1
                                                                     July 1992

-------
      organochlorine   pesticides   and    PCBs,   nitroaromatics,   haloethers,
      chlorinated hydrocarbons,  and organophosphorus  pesticides.   Just before
      use, activate  each  batch at least  16  nr at 130°C  in  a  glass container
      loosely covered with aluminum foil.  Alternatively,  store the Florisil in
      an oven at 130°C.  Cool  the Florisil  before use in a desiccator.  (Florisil
      from different  batches  or sources may vary in  adsorptive  capacity.   To
      standardize the amount of Florisil which is used, the use of lauric acid
      value is  suggested.  The referenced procedure  determines the adsorption
      from hexane solution of lauric acid  (mg) per g of Florisil.  The amount of
      Florisil to be used for each column  is  calculated by dividing 110 by this
      ratio and multiplying by 20 g (Mills).

      5.3   Sodium sulfate (granular,  anhydrous), Na2S04 - Purify by heating at
400°C for 4 hours in a shallow tray,  or by precleaning the sodium sulfate with
methylene chloride.   If the sodium sulfate  is precleaned with methylene chloride,
a method blank must be analyzed, demonstrating that there is no interference from
the sodium sulfate.

      5.4   Eluting solvents

            5.4.1 Diethyl  ether,  C2H5OC^H5 - Pesticide  quality  or equivalent.
      Must be free  of  peroxides, as  indicated  by test  strips (EM  Quant or
      equivalent).  Procedures recommended for removal  of peroxides are provided
      with the  test  strips.   After cleanup, 20  ml  ethyl  alcohol preservative
      must be added to each liter of ether.

            5.4.2 Acetone, CH3COCH3 -  Pesticide  quality or equivalent.

            5.4.3 Hexane, C6H14  -  Pesticide quality or  equivalent.

            5.4.4 Methylene chloride,  CH2C12 - Pesticide quality or equivalent.

            5.4.5 Pentane, CH3(CH2)3CH3 - Pesticide quality or equivalent.

            5.4.6 Petroleum ether (boiling range 30-60°C) - Pesticide quality or
      equivalent.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory material to this chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Phthalate esters

            7.1.1 Reduce the  sample  extract volume to 2  mL  prior to cleanup.
      The extract solvent must be hexane.

            7.1.2 Place  approximately  10 g  of deactivated  Florisil  (Section
      5.1.1) into a 10 mm ID chromatographic column.  Tap the column to settle
      the Florisil and add approximately  1 cm of anhydrous sodium sulfate to the

                                   3620A  - 3                        Revision 1
                                                                     July 1992

-------
top.

      7.1.3 Preelute the column with 40 ml of  hexane.   The  rate for all
elutions should be about 2  mL/min.  Discard the eluate and, just prior to
exposure of the sodium sulfate layer to the air,  quantitatively transfer
the 2 ml sample extract onto the column using  an additional 2 ml of hexane
to complete the transfer.  Just prior  to  exposure  of the  sodium sulfate
layer to the  air,  add 40 ml  of hexane and continue the  elution  of the
column.  Discard this hexane eluate.

      7.1.4 Next,  elute  the  column with  100 ml of  20% ethyl  ether in
hexane  (v/v)  into  a  500 ml K-D flask  equipped  with a 10 ml  concentrator
tube.  Concentrate the collected fraction  as needed.  No solvent exchange
is necessary.   Adjust the  volume  of the  cleaned-up  extract  to whatever
volume  is  required  (10  ml  for  Method  8060)  and  analyze  by  gas
chromatography.  Compounds that elute in this fraction are:

      Bis(2-ethylhexyl) phthalate
      Butyl benzyl phthalate
      Di-n-butyl phthalate
      Diethyl phthalate
      Dimethyl phthalate
      Di-n-octyl phthalate

7.2   Nitrosamines

      7.2.1 Reduce the sample extract volume to 2 ml prior to cleanup.

      7.2.2 Add  a   weight   of  activated  Florisil   (nominally  22  g)
predetermined   by   calibration   (Section  5.1.2)   into  a  20   mm  ID
chromatographic column.  Tap the  column to settle the  Florisil  and add
about 5 mm of anhydrous sodium sulfate to the top.

      7.2.3 Pre-elute the column with 40 ml of ethyl ether/pentane (15:85)
(v/v).   Discard the  eluate  and,  just prior to  exposure of  the  sodium
sulfate layer to the air,  quantitatively  transfer the 2 ml sample extract
onto  the  column using  an  additional  2  ml of  pentane to  complete the
transfer.

      7.2.4 Elute  the  column with 90 ml  of  ethyl  ether/pentane (15:85)
(v/v)   and  discard   the   eluate.     This  fraction  will  contain  the
diphenylamine,  if  it  is present in the extract.

      7.2.5 Next,  elute  the  column with  100 ml of  acetone/ethyl  ether
(5:95)  (v/v)  into  a  500 ml K-D  flask equipped  with a 10 ml concentrator
tube.   This  fraction will  contain  all  of the nitrosamines listed in the
scope of the method.

      7.2.6 Add 15 ml of methanol  to the collected fraction, concentrate
as needed using pentane to  prewet the K-D column and set the water bath at
70 to 75°C.  When the apparatus is cool, remove the Snyder column and rinse
the flask and its lower joint into the concentrator tube with 1  to  2 ml of
pentane.


                             3620A -  4                         Revision 1
                                                               July  1992

-------
      7.3   Organochlorine   pesticides,   haloethers,   and   organophosphorus
pesticides (see Tables 1  and 2  for fractionation patterns of compounds tested)

            7.3.1 Reduce the sample extract volume to  10  ml  prior to cleanup.
      The extract solvent must be hexane.

            7.3.2 Add  a  weight  of  activated  Florisil   (nominally  20  g),
      predetermined   by   calibration   (Section   5.1.2),   to   a   20  mm   ID
      chromatographic column.  Settle  the Florisil by tapping the  column.  Add
      anhydrous sodium sulfate to the  top of the Florisil  to  form  a layer 1 to
      2 cm deep.  Add 60 ml  of hexane  to wet  and  rinse the sodium sulfate  and
      Florisil.  Just prior to exposure of the sodium sulfate to air, stop  the
      elution  of  the hexane by closing  the  stopcock  on  the  chromatographic
      column.  Discard the eluate.

            7.3.3 Adjust the sample extract volume  to 10 ml with  hexane  and
      transfer it from the  K-D  concentrator  tube to the Florisil column.  Rinse
      the tube twice with 1-2 ml hexane,  adding each  rinse to the  column.

            7.3.4 Place a 500 ml K-D flask and clean concentrator tube under the
      chromatographic column.  Drain the  column  into  the flask until the sodium
      sulfate layer is nearly exposed.  Elute the column with  200 ml of 6% ethyl
      ether in hexane (v/v)  (Fraction  1) using  a  drip  rate of about 5 mL/min.
      All of the haloethers are in this fraction.  Remove the K-D flask and  set
      aside for later concentration.   Elute the column again,  using 200  ml of
      15% ethyl ether in hexane (v/v)  (Fraction 2),  into  a  second  K-D flask.
      Perform a third elution  using 200  ml  of  50% ethyl ether in  hexane  (v/v)
      (Fraction  3),  and  a  final  elution   with  200  ml  of  100%  ethyl  ether
      (Fraction 4), into separate K-D  flasks.

            7.3.5 If  necessary,  concentrate  the eluates  by standard  K-D
      techniques  using  the water bath at  about  85°C  (75°C  for  Fraction  4).
      Adjust the final volume to whatever volume is required  (1-10 ml).

      7.4   Nitroaromatics  and  isophorone

            7.4.1 Reduce the sample  extract  volume to 2 ml prior to cleanup.

            7.4.2 Add  a  weight  of   activated   Florisil  (nominally   10   g)
      predetermined  by  calibration   (Section  5.1.2)   into   a   10  mm   ID
      chromatographic column.   Tap the column to settle the  Florisil  and  add
      about 1 cm of anhydrous sodium sulfate to the top.

            7.4.3 Pre-elute  the column  with  methylene chloride/hexane  (1:9)
      (v/v) at about 2 mL/min.   Discard the  eluate and, just  prior to exposure
      of the sodium sulfate layer to  the air, quantitatively transfer the sample
      extract onto the column using an  additional 2 ml of hexane to complete  the
      transfer.  Just prior to  exposure of the sodium sulfate layer to the air,
      add  30  ml  of  methylene   chloride/hexane  (1:9)  (v/v)  and  continue  the
      elution of the column.  Discard  the eluate.

            7.4.4 Elute the  column with  90  ml of ethyl ether/pentane  (15:85)
      (v/v)  and  discard  the  eluate.     This   fraction  will   contain  the
      diphenylamine,  if it  is present  in  the extract.

                                  3620A  - 5                         Revision  1
                                                                     July 1992

-------
      7.4.5 Next, elute  the  column with  100  mL of  acetone/ethyl  ether
(5:95) (v/v) into a 500 ml K-D  flask  equipped  with  a 10 ml concentrator
tube,  this fraction will contain  all  of  the  nitrosamines  listed in the
scope of the method.

      7.4.6 Add 15 mL of methanol to the collected fraction, concentrate
using pentane to prewet the K-D  column, and set  the  water  bath  at  70 to
75°C.   When the  apparatus  is cool,  remove the Snyder column and rinse the
flask and  its lower joint into  the concentrator tube with  1  to  2  ml of
pentane.

      7.4.7 Next,  elute  the  column  with  30 ml  of  acetone/methylene
chloride  (1:9)  (v/v)   into  a 500  ml  K-D  flask  equipped  with  a  10 ml
concentrator tube.  Concentrate the collected  fraction,  while exchanging
the  solvent to  hexane.   To  exchange the  solvent,  reduce  the  elution
solvent to  about 10 mL.   Add  50  mL of hexane,  a  fresh boiling chip, and
return the  reassembled K-D apparatus  to the hot  water bath.  Adjust the
final volume of the cleaned-up extract to whatever volume is required (1-
10 mL).  Compounds that elute in this fraction are:

      2,4-Dinitrotoluene
      2,6-Dinitrotoluene
      Isophorone
      Nitrobenzene.

7.5   Chlorinated hydrocarbons

      7.5.1 Reduce the  sample extract volume  to 2 mL  prior to cleanup.
The extract solvent must be hexane.

      7.5.2 Add  a  weight   of   activated   Florisil   (nominally  12  g)
predetermined   by   calibration   (Section  5.1.2)   into  a   10  mm  ID
chromatographic  column.   Tap the column to settle  the  Florisil  and add
about 1 to  2 cm of anhydrous  sodium sulfate to the top.

      7.5.3 Preelute the column with  100 mL of petroleum ether.  Discard
the eluate  and, just prior to exposure of  the  sodium  sulfate layer to the
air,  quantitatively  transfer  the  sample extract   to the  column  by
decantation and subsequent petroleum ether washings.  Discard  the eluate.
Just  prior to exposure  of  the  sodium  sulfate layer to the  air,   begin
eluting the column with 200 mL of  petroleum ether and collect the eluate
in  a  500  mL K-D  flask equipped with  a  10 mL concentrator tube.   This
fraction should  contain all of the chlorinated hydrocarbons:

      2-Chloronaphthalene
       1,2-Dichlorobenzene
       1,3-Dichlorobenzene
       1,4-Dichlorobenzene
      Hexachlorobenzene
      Hexachlorobutadiene
      Hexachlorocyclopentadi ene
      Hexachloroethane
       1,2,4-Tri chlorobenzene.


                             3620A - 6                        Revision  1
                                                               July 1992

-------
            7.5.4 Concentrate the fraction, using hexane to prewet the column.
      When the apparatus is cool, remove the Snyder column and rinse the flask
      and its lower joint  into  the  concentrator  tube  with hexane.   Adjust the
      final volume  of the cleaned-up  extract  to whatever volume  is required
      (1-10 ml).


8.0   QUALITY CONTROL

      8.1   Refer to  Chapter  One for specific quality  control  procedures and
Method 3600 for cleanup procedures.

      8.2   The analyst should  demonstrate  that  the  compounds of interest are
being quantitatively recovered before applying this method to actual samples.

      8.3   For sample  extracts that  are  cleaned up  using  this  method,  the
associated quality control  samples should also be  processed through this cleanup
method.
9.0   METHOD PERFORMANCE

      9.1   Table 1 indicates the distribution of chlorinated pesticides, PCB's,
and haloethers in various Florisil column fractions.

      9.2   Table 2 indicates the distribution of organophosphorus pesticides in
various Florisil  column fractions.
10.0  REFERENCES

1.    Gordon, A.J.  and  R.A.  Ford, The  Chemist's Companion:   A  Handbook of
      Practical  Data, Techniques,  and  References  (New York:  John Wiley & Sons,
      Inc.), pp. 372, 374, and 375, 1972.

2.    Floridin of ITT System,  Florisil:   Properties, Application, Bibliography,
      Pittsburgh, Pennsylvania, 5M381DW.

3.    Mills, P.A., "Variation  of Florisil Activity; Simple Method for Measuring
      Absorbent Capacity and its use in Standardizing  Florisil Columns," Journal
      of the Association of Official Analytical Chemists, 51, 29, 1968.

4.    U.S. Food and Drug Association,  Pesticides Analytical Manual (Volume 1),
      July 1985.

5.    U.S. EPA 40 CFR Part 136, "Guidelines  Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final Rule and Interim
      Final Rule and Proposed Rule," October 26,  1984.
                                   3620A  -  7                         Revision 1
                                                                     July 1992

-------
                             TABLE  1

          DISTRIBUTION OF CHLORINATED  PESTICIDES.  PCBs.
          AND HALOETHERS  INTO  FLORISIL COLUMN  FRACTIONS
Parameter
                                       Percent Recovery by Fraction8
                    1
Aldrin
a-BHC
B-BHC
y-BHC
5-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Haloethers
Heptachlor
Heptachlor epoxide
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
100
100
97
98
100
100
99
98
100
0 100
37 64
0 7
0 0
4 96
0 68
R
100
100
96
97
97
95 4
97
103
90
95











91
106

26











   Eluant composition:
Fraction 1 -  6% ethyl  ether in hexane
Fraction 2 - 15% ethyl  ether in hexane
Fraction 3 - 50% ethyl  ether in hexane
R = Recovered  (no percent recovery data presented).

SOURCE:  U.S.  EPA and FDA data.
                             3620A - 8
                                      Revision 1
                                       July 1992

-------
R
NR
V
ND
                                   TABLE  2

                  DISTRIBUTION  OF ORGANOPHOSPHORUS  PESTICIDES
                        INTO FLORISIL COLUMN FRACTIONS
                                      Percent Recovery by Fraction8
Parameter
Azinphos methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Monochrotophos
Naled
Parathion
Parathion methyl
Phorate
Ronnel
1

ND
>80
NR
100

NR
ND
25-40

V
ND
R

V
ND
ND
NR


0-62
>80
Stirophos (Tetrachlorvinphos) ND
Sulfotepp
TEPP
Tokuthion (Prothiofos)
Trichloronate
Eluant composition:



V
ND
>80
>80
Fraction 1 -
Fraction 2 -
Fraction 3 -
Fraction 4 -
2

ND

NR

100
NR
ND

>80
V
ND
R
5
V
ND
ND
NR
100
100


ND
V
ND


200 mL
200 ml
200 ml
200 mL
3 4
20 80
ND ND

NR


NR
ND ND


V
ND ND

95
V
ND ND
ND ND
NR




ND ND

ND ND


of 6% ethyl ether in
of 15% ethyl ether in
of 50% ethyl ether in
of 100% ethyl ether




























hexane
hexane
hexane

Recovered (no percent recovery information presented) (U.S. FDA)
Not recovered (U.S. FDA).
Variable recovery (U.S. FDA).
Not determined.
SOURCE:  U.S. EPA and FDA data.
                                   3620A -  9
                                                        Revision 1
                                                         July 1992

-------
                                        METHOD  3620A
                                 FLORISIL COLUMN CLEANUP
711  Reduce volume
 of sample extract
      to  2 mL
                      Phthalate Esters
    7  1  2 Place
   Florisil  into
  chromatographic
    column;  add
 anhydrous sodium
      sulfate
  7  1  3  Preelute
column with hexane,
  transfer sample
extract, add hexane
7 1  4  Elute column
with ethyl ether in
      hexane
                                                                           721 Reduce volume
                                                                            of sample extract
                                                                                 to 2 mL
                                                                           7 2 2 Put Florisil
                                                                                  into
                                                                             chromatographic
                                                                               column;  add
                                                                            anhydrous sodium
                                                                                 sulfate
                                                                             7 2 3 Preelute
                                                                            column with ethyl
                                                                             ether/pentane,
                                                                            transfer extract,
                                                                               add pentane
                                                                            7 2 4 Elute column
                                                                               with ethyl
                                                                              ether/pentane
 7 1 4  Concentrate
 fraction, adjust
  volume, analyze
                                                                            7 2 5 Elute column
                                                                            with acetone/ethyl
                                                                            ether into flask
                                          3620A -  10
                                                                                       Revision  1
                                                                                        July  1992

-------
                      METHOD  3620A
                         continued
Organochl or me
pesticides, haloethers
and organophosphorous
731 Reduce volume
 of  sample ex tract
     to 2 mL
  troa romatics
and  is ophorone
                        Chi orinated
                        hydrocarbons
                 Ana 1yze by CC
                                      741  Reduce volume
                                       of sample ex tract
                                            to  2 mL
Fl onail
tographic
n, add
* sodium
e then
discard
ate


ust sample
volume ,
to column
Lh hexane



in col umn ,
olumn 4
o separate
iks


icent ra te
ad jus t
jme























7 S 1 Reduce volume
of sample ex tract
to 2 mL



7 5 2



Place
Florisil in
chroma tographic
col urn. add
anhydrous s oaium
sulfate

7 5 3 P

*eel u te
col umn with
pet r o leum ether ,
t ransf er sampl e
ex t rac t to col umn ,
discard

ei -a te

7 5 4 Concentrate
i rac tion
aQ jus t
final vo 1 ume

•~\





















7 4 2 Put
3 lur t
chroma tc
col umr
anhydrous
sul;


7 4 3 Tt
sample ext
col umn
methy
chl o r ide/
discard


7 4 4 Elut
wit
ace t one/m
chloride
solvent t

7 4 4 Con
f racti on ,
fina 1 v


                                      2 6 Add melhanol
                                       to f raction;
                                        concentrate
                                                 add
                                               anafer
                                               .ract on
                                               i, add
                                               excnange
                       3620A -  11
                                             Revision  1
                                              July  1992

-------
                                 METHOD 3630B

                              SILICA GEL CLEANUP
1.0   SCOPE AND APPLICATION

      1.1   Silica gel  is a  regenerative  adsorbent of amorphous  silica with
weakly acidic properties.  It is produced from sodium silicate and sulfuric acid.
Silica gel can be used in column chromatography for the separation of analytes
from interfering  compounds  of a different chemical polarity.   It  may be used
activated, after heating to 150 - 160°C,  or deactivated with up to  10% water.

      1.2   This method includes guidance  for standard  column cleanup of sample
extracts  containing  polynuclear  aromatic  hydrocarbons,  derivatized  phenolic
compounds, organochlorine pesticides, and PCBs as Aroclors.

      1.3   This  method  also  provides  cleanup  procedures  using  solid-phase
extraction  cartridges  for   pentafluorobenzyl   bromide-derivatized  phenols,
organochlorine pesticides, and PCBs as Aroclors.  This technique also provides
the best  separation of PCBs from most single component organochlorine pesticides.
When only PCBs are to be measured, this method can be used in conjunction with
sulfuric acid/permanganate cleanup (Method 3665).

      1.4   Other analytes may be cleaned  up using this method if  the analyte
recovery meets the criteria specified in Sec.  8.0.


2.0   SUMMARY OF METHOD

      2.1   This  method provides  the option of using  either  standard column
chromatography techniques  or solid-phase extraction cartridges.  Generally, the
standard column chromatography techniques use larger amounts of adsorbent and,
therefore, have a greater cleanup capacity.

      2.2   In the standard column cleanup protocol, the column is  packed with
the required amount of adsorbent, topped with a water adsorbent, and then loaded
with the sample to be analyzed.  Elution of the analytes is accomplished with a
suitable sol vent(s) that leaves the  interfering  compounds  on  the column.   The
eluate is then concentrated (if necessary).

      2.3   The cartridge cleanup protocol  uses  silica solid-phase extraction
cartridges packed with 1 g or 2 g of adsorbent.   Each cartridge is solvent washed
immediately prior  to use.   Aliquots of sample  extracts  are  loaded  onto  the
cartridges, which are then eluted with suitable solvent(s).  A vacuum manifold
is required  to obtain reproducible  results.   The collected fractions  may  be
further concentrated  prior to gas chromatographic analysis.

      2.4   The appropriate gas chromatographic method is  listed at the end of
each technique.   Analysis  may also  be  performed by  gas  chromatography/mass
spectrometry (Method  8270).
                                  3630B  -  1                         Revision 2
                                                                September 1994

-------
3.0   INTERFERENCES

      3.1   Solvents, reagents, glassware,  and other sample processing hardware
may yield discrete  artifacts and/or elevated baselines causing misinterpretation
of gas chromatograms.  All  these materials  must be demonstrated to be free from
interferences under the conditions of the analysis, by analyzing reagent blanks.
See Sec. 8 for guidance on a reagent blank check.

      3.2   Phthalate  ester  contamination  may  be  a  problem  with  certain
cartridges  The more inert  the  column and/or  cartridge material (i.e., glass or
Teflon),  the  less  problem  with phthalates.   Phthalates create  interference
problems for all method analytes, not just the phthalate esters themselves.

      3.3   More extensive procedures than  those  outlined  in this method may be
necessary for reagent purification.


4.0   APPARATUS AND MATERIALS

      4.1   Chromatographic column - 250 mm  long x 10  mm  ID;  with Pyrex glass
wool at bottom and a Teflon stopcock.

      NOTE: Fritted  glass  discs are difficult  to  decontaminate  after  highly
            contaminated extracts  have  been passed  through.   Columns  without
            frits  may  be  purchased.   Use  a  small pad  of Pyrex glass  wool to
            retain  the adsorbent.   Prewash  the  glass  wool  pad with 50  ml of
            acetone followed by  50 ml of elution solvent  prior to packing the
            column with adsorbent.

      4.2   Beakers - 500 ml.

      4.3   Vials  - 2, 10, 25  ml,  glass with Teflon  lined screw-caps or crimp
tops.

      4.4   Muffle furnace.

      4.5   Reagent bottle - 500 ml.

      4.6   Erlenmeyer flasks  - 50 and  250 ml.

      4.7   Vacuum   manifold:    VacElute    Manifold   SPS-24   (Analytichem
International),  Visiprep  (Supelco, Inc.)  or equivalent,  consisting of glass
vacuum basin, collection rack and funnel, collection vials,  replaceable stainless
steel delivery  tips, built-in  vacuum  bleed valve  and  gauge.   The  system is
connected to a vacuum pump  or water aspirator through a vacuum trap made from a
500 ml sidearm flask fitted with a one-hole  stopper  and glass tubing.
5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be used in all  tests.  Unless
otherwise  indicated,  it is  intended  that all  reagents  shall  conform  to  the
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where  such  specifications  are available.   Other grades  may be used,

                                  3630B  -  2                         Revision 2
                                                                September 1994

-------
provided it is first  ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.  All  references to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Silica gel for chromatography columns.

            5.3.1 Silica Gel  for Phenols and  Polynuclear Aromatic Hydrocarbons:
      100/200 mesh desiccant  (Davison Chemical grade 923 or equivalent).  Before
      use, activate for  at least 16 hr.  at 130°C in a shallow glass tray, loosely
      covered with foil.

            5.3.2 Silica Gel  for Organochlorine pesticides/PCBs:   100/200 mesh
      desiccant  (Davison Chemical  grade  923 or  equivalent).    Before  use,
      activate for at least  16 hr. at  130°C  in a  shallow glass tray, loosely
      covered with foil.  Deactivate it to 3.3% with reagent water in a 500 ml
      glass jar.   Mix the  contents thoroughly and  allow  to  equilibrate for 6
      hours.  Store  the  deactivated silica gel  in  a sealed glass  jar inside a
      desiccator.

      5.4   Silica cartridges:  40 fj,m particles, 60 A pores. The cartridges with
which this method was developed consist of 6  mL serological-grade polypropylene
tubes, with the 1 g  of silica held between two polyethylene or stainless steel
frits with 20 /xm pores.   2 g  silica cartridges are also used in  this method, and
0.5 g cartridges are  available.   The compound elution  patterns  must be verified
when cartridges other than the specified size are used.

      5.5   Sodium sulfate (granular,   anhydrous), Na2S04.  Purify by heating at
400°C for 4 hours in  a shallow tray,  or by precleaning the sodium  sulfate with
methylene chloride.  A method blank must be analyzed in order to  demonstrate that
there is no interference from the sodium sulfate.

      5.6   Eluting  solvents

            5.6.1 Cyclohexane,  C6H12 -  Pesticide quality  or equivalent.

            5.6.2 Hexane, C6H14  - Pesticide quality  or  equivalent.

            5.6.3 2-Propanol, (CH3)2CHOH - Pesticide quality  or equivalent.

            5.6.4 Toluene,  C6H5CH3  - Pesticide quality  or  equivalent.

            5.6.5 Methylene chloride,  CH2C12  - Pesticide quality or equivalent.

            5.6.6 Pentane,  C5H12 -  Pesticide  quality or equivalent.

            5.6.7 Acetone,  CH3COCH3 - Pesticide quality or equivalent.

            5.6.8 Diethyl Ether, C2H5OC2H5.    Pesticide  quality or  equivalent.
      Must  be  free  of  peroxides  as indicated  by  test  strips (EM Quant,  or
      equivalent).   Procedures for removal of peroxides  are  provided with the
                                   3630B  -  3                         Revision 2
                                                                September 1994

-------
      test strips.  After cleanup, 20 ml of ethanol  preservative must be added
      to each liter of ether.

6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material to this chapter,  Organic Analytes, Sec.
4.1.


7.0   PROCEDURE

      7.1   General Guidance

            7.1.1 The procedure contains two cleanup  options for the derivatized
      phenols and organochlorine pesticides/Aroclors, but  only one technique for
      the   polynuclear   aromatic   hydrocarbons    (PAHs)   (standard   column
      chromatography).   Cleanup techniques by standard column chromatography for
      all analytes are  found in  Sec. 7.2.   Cleanup techniques  by  solid-phase
      cartridges for derivatized  phenols  and PAHs are found in  Sec.  7.3.   The
      standard column chromatography  techniques are packed with a greater amount
      of silica gel adsorbent and, therefore, have a greater  cleanup capacity.
      A  rule  of thumb  relating  to  cleanup  capacity  is  that  1  g  of  sorbent
      material  will  remove  10  to 30  mg  of total  interferences.    (However,
      capacity  is  also  dependent   on   the  sorbent  retentiveness  of  the
      interferences.)   Therefore, samples  that  exhibit a  greater  degree  of
      sample interference should be cleaned up by the standard column technique.
      However, both techniques have  limits on  the amount of  interference that
      can be removed.  If the interference is caused by  high  boiling material,
      then Method 3640  should be used prior to this method.  If the interference
      is caused by relatively polar compounds of the same boiling range as the
      analytes, then  multiple column  or cartridge cleanups may be required.  If
      crystals of sulfur are noted in  the extract,  then  Method  3660 should be
      utilized prior  to this  method.  The cartridge cleanup techniques are often
      faster and use  less solvent, however they have less cleanup capacity.

            7.1.2 Allow the extract to reach  room temperature if it was in cold
      storage.   Inspect the extracts visually to  ensure that  there  are  no
      particulates or phase separations and that the volume is as stated in the
      accompanying documents.  Verify  that the solvent is compatible with the
      cleanup procedures.  If crystals of  sulfur are visible or if the presence
      of sulfur is suspected, proceed with Method  3660.

            7.1.3 If the extract solvent is methylene chloride, for most cleanup
      techniques, it must be exchanged to hexane.   (For  the  PAHs,  exchange to
      cyclohexane  as  per  Sec. 7.2.1).    Follow  the  standard  Kuderna-Danish
      concentration technique provided in  each extraction  method.  The volume of
      methylene chloride should have been reduced to  1  - 2 mL.  Add 40 ml of
      hexane, a fresh boiling chip and repeat the concentration as written.  The
      final volume required  for the cleanup techniques is normally 2 ml.
                                   3630B  -  4                         Revision 2
                                                                September 1994

-------
7.2   Standard Column Cleanup Techniques

      7.2.1 Polynuclear aromatic hydrocarbons

            7.2.1.1     Before the  silica  gel  cleanup technique  can  be
      utilized,  the extract solvent  must be exchanged to cyclohexane. The
      exchange  is  performed  by  adding 4  ml  of cyclohexane  following
      reduction  of the  sample extract  to  1-2 mL using the  macro  Snyder
      column.    Attach  the two ball micro  Snyder column and  reduce the
      volume to  2 mL.

            CAUTION:     When the volume of  solvent is reduced below 1 ml,
                        semivolatile  analytes   may  be  lost.    If  the
                        extract goes to dryness, the  extraction must  be
                        repeated.

            7.2.1.2     Prepare a slurry of 10  g of  activated silica gel
      (Sec. 5.3.1)  in methylene chloride and place  this  into  a  10 mm  ID
      chromatographic column.  Tap the column to settle the silica gel and
      elute the  methylene  chloride.   Add  1 to  2 cm of anhydrous  sodium
      sulfate  to the top of the silica gel.

            7.2.1.3     Pre-elute the  column with 40 ml  of  pentane.   The
      rate for all  elutions should be about 2 mL/min.  Discard the  eluate
      and, just  prior to exposure of the sodium sulfate layer to the air,
      transfer the  2 ml cyclohexane  sample  extract onto  the  column using
      an additional 2 mL cyclohexane to complete the  transfer.  Just prior
      to exposure of the  sodium  sulfate layer  to the  air,   add 25  ml  of
      pentane  and  continue the  elution of  the column.    Discard  this
      pentane  eluate.

            7.2.1.4     Next,  elute the column  with 25  mL  of  methylene
      chloride/pentane (2:3)(v/v)  into  a 500 mL K-D flask equipped  with a
      10 mL concentrator  tube.   Concentrate the collected  fraction  to
      whatever volume is required (1-10 mL).   Proceed with  HPLC  (Method
      8310) or GC analysis (Method 8100).  Validated components that elute
      in this  fraction are:

            Acenaphthene
            Acenaphthylene
            Anthracene
            Benzo(a)anthracene
            Benzo(a)pyrene
            Benzo(b)fluoranthene
            Benzo(g,h,i)pery1ene
            Benzo(k)fluoranthene
            Chrysene
            Dibenzo(a,h)anthracene
            Fluoranthene
            Fluorene
            Indeno(l,2,3-cd)pyrene
            Naphthalene
            Phenanthrene
            Pyrene

                            3630B - 5                         Revision 2
                                                         September  1994

-------
7.2.2  Derivatized Phenols

      7.2.2.1     This silica gel  cleanup procedure is performed on
sample  extracts  that  have   undergone  pentafluorobenzyl  bromide
derivatization, as described  in  Method 8040.  The  sample extract
must be in 2 ml of hexane at this point.

      7.2.2.2     Place 4.0 g of activated silica gel  (Sec. 5.3.1)
into a 10 mm  ID chromatographic  column.   Tap the column to settle
the silica gel and add about  2 g of anhydrous sodium sulfate to the
top of the silica gel.

      7.2.2.3     Pre-elute the column with  6 mL of  hexane.   The
rate for all  elutions should  be about 2 mL/min.  Discard the eluate
and, just prior to exposure of the sodium sulfate layer to the air,
pipet onto the column 2 mL of the hexane solution that contains the
derivatized sample or  standard.  Elute the column with  10.0 mL of
hexane and discard the eluate.

      7.2.2.4     Elute the column, in order, with  10.0 ml of 15%
toluene in hexane  (Fraction  1);  10.0  mL of  40%  toluene in hexane
(Fraction 2);  10.0 mL of 75% toluene  in hexane  (Fraction 3); and
10.0 mL  of  15% 2-propanol in toluene  (Fraction  4).  All elution
mixtures are  prepared  on  a volume:volume  basis.   Elution patterns
for the phenolic derivatives  are shown in Table 1.  Fractions may be
combined,  as  desired,  depending  upon  the   specific  phenols  of
interest  or   level  of  interferences.    Proceed  with  GC  analysis
(Method 8040).

7.2.3 Organochlorine Pesticides and Aroclors

      7.2.3.1     Transfer a 3 g portion of deactivated  silica gel
(Sec. 5.3.2)  into  a 10 mm  ID glass chromatographic column and top it
with 2 to 3 cm of anhydrous sodium sulfate.

      7.2.3.2     Add  10  mL  of hexane  to  the top of the column to
wet  and  rinse the sodium  sulfate and  silica gel.   Just  prior to
exposure of the sodium sulfate layer to air,  stop the hexane eluate
flow by closing the stopcock on the chromatographic column.  Discard
the eluate.

      7.2.3.3     Transfer the sample extract (2  mL in hexane) onto
the column. Rinse the  extract vial twice  with  1  to  2  mL of hexane
and add each  rinse to  the column.  Elute  the column with 80 mL of
hexane  (Fraction  I)  at  a rate  of about  5  mL/min.    Remove the
collection flask  and  set  it  aside  for  later  concentration.  Elute
the  column  with  50  mL  of hexane (Fraction  II) and collect the
eluate.  Perform  a third  elution  with  15  mL  of methylene chloride
(Fraction  III).  The  elution  patterns  for  the  organochlorine
pesticides,  Aroclor-1016,  and Aroclor-1260 are shown in  Table 2.

      7.2.3.4     Prior   to   gas   chromatographic   analysis,   the
extraction solvent must be exchanged  to hexane.  Fractions may be
combined,    as    desired,   depending    upon    the    specific

                       3630B  - 6                         Revision 2
                                                    September 1994

-------
      pesticides/Aroclors  of  interest or  level  of  interferences.   If
      mixtures of  Aroclors and pesticides  are  expected,  it  is  best to
      analyze  Fraction   I  separately, since  it contains  the  Aroclors
      separated from  most  pesticides. Proceed with  GC analysis  as  per
      Method 8081.

7.3   Cartridge Cleanup Techniques

      7.3.1 Cartridge Set-up and Conditioning

            7.3.1.1     Arrange the 1  g silica cartridges (2 g for phenol
      cleanup) on the manifold in the  closed-valve  position.   Other size
      cartridges may be  used, however the  data presented in the Tables are
      all  based  on  1  g  cartridges  for  pesticides/Aroclors  and  2  g
      cartridges for phenols.  Therefore,  supporting recovery data must be
      developed for other sizes.   Larger cartridges will probably require
      larger volumes of elution solvents.

            7.3.1.2     Turn on the vacuum  pump and set pump vacuum to 10
      inches   (254  mm)  of  Hg.   Do   not   exceed   the  manufacturer's
      recommendation for manifold  vacuum. Flow rates can be controlled by
      opening and closing cartridge valves.

            7.3.1.3     Condition the  cartridges  by adding 4 ml of hexane
      to each cartridge.  Slowly open the  cartridge  valves to allow hexane
      to pass through the  sorbent  beds to  the lower  frits.   Allow a  few
      drops per cartridge to pass through the manifold to remove all  air
      bubbles.  Close the valves and allow  the solvent to soak the entire
      sorbent bed for 5 minutes.   Do not  turn off the vacuum.

            7.3.1.4     Slowly open cartridge valves  to allow the hexane
      to pass  through the  cartridges.   Close the  cartridge  valves when
      there is still at least 1 mm of  solvent above the sorbent bed.   Do
      not allow cartridges to become  dry.   If cartridges  go  dry,  repeat
      the conditioning step.

      7.3.2 Derivatized Phenols

            7.3.2.1     Reduce the sample extract volume to 2 ml prior to
      cleanup.  The extract solvent must be  hexane and the  phenols must
      have undergone derivatization by pentafluorobenzyl bromide,  as  per
      Method 8040.

            7.3.2.2     Transfer the extract to  the  2  g  cartridge that has
      been conditioned as  described in Sec.  7.3.1.  Open the cartridge
      valve to allow  the extract  to  pass  through  the cartridge  bed at
      approximately 2 mL/minute.

            7.3.2.3     When the  entire  extract  has  passed  through  the
      cartridges,  but before the cartridge  becomes  dry, rinse the sample
      vials with  an additional  0.5 ml  of hexane,  and add the rinse to  the
      cartridges  to complete the quantitative transfer.
                            3630B - 7                         Revision 2
                                                          September 1994

-------
      7.3.2.4     Close the cartridge  valve  and turn off the vacuum
after the solvent has  passed  through,  ensuring  that  the cartridge
never gets dry.

      7.3.2.5     Place a  5  ml  vial  or volumetric flask  into the
sample  rack  corresponding to  the cartridge  position.    Attach  a
solvent-rinsed stainless steel solvent guide to the manifold cover
and align with the collection vial.

      7.3.2.6     Add 5 mL of hexane to the  cartridge.  Turn on the
vacuum pump and adjust the pump  pressure  to 10  inches  (254 mm) of
Hg.  Allow the  solvent  to soak the sorbent  bed for 1 minute or less.
Slowly open the cartridge  valve, and  collect the  eluate  (this is
Fraction 1,  and should be discarded).

      NOTE:  If cartridges smaller than 2  g are used,  then Fraction
            1  cannot be  discarded,  since  it contains  some  of the
            phenols.

      7.3.2.7     Close the cartridge  valve,  replace the collection
vial, and add 5 ml of toluene/hexane (25/75,  v/v) to the cartridge.
Slowly open  the  cartridge valve  and  collect the  eluate  into the
collection vial.  This is  Fraction 2, and  should  be  retained for
analysis.

      7.3.2.8     Adjust the final volume  of the eluant to a known
volume which will  result in analyte concentrations  appropriate for
the  project  requirements  (normally  1-10 ml).   Table  3  shows
compound  recoveries  for 2  g  silica  cartridges.   The  cleaned up
extracts are ready for analysis by Method  8040.

7.3.3 Organochlorine Pesticides/Aroclors

NOTE: The   silica   cartridge   procedure    is  appropriate   when
      polychlorinated biphenyls are known  to be present.

      7.3.3.1     Reduce the sample extract  volume  to 2 ml prior to
cleanup.  The extract solvent must be  hexane.

      7.3.3.2     Use the 1 g cartridges conditioned as described in
Sec. 7.3.1.

      7.3.3.3     Transfer the extract to  the cartridge.  Open the
cartridge valve to allow the extract to pass through the cartridge
bed at approximately 2 mL/minute.

      7.3.3.4     When the  entire extract has passed  through the
cartridges,  but before the cartridge becomes dry, rinse the sample
vials with an additional  0.5 ml of solvent, and add  the rinse to the
cartridges to complete the quantitative transfer.

      7.3.3.5     Close the cartridge  valve  and turn off the vacuum
after the solvent has  passed  through,  ensuring  that  the cartridge
never goes dry.

                       3630B  -  8                         Revision 2
                                                    September 1994

-------
                  7.3.3.6     Place a  5  ml vial or  volumetric  flask into the
            sample  rack corresponding to  the  cartridge  position.    Attach  a
            solvent-rinsed stainless steel solvent guide to the manifold cover
            and align with the collection vial.

                  7.3.3.7     Add 5 mL of hexane to  the  cartridge.  Turn on the
            vacuum pump and  adjust  the pump  pressure to 10 inches (254 mm) of
            Hg.  Allow the solvent to soak the sorbent bed for  1 minute or less.
            Slowly  open  the  cartridge valve  and collect the eluate  into the
            collection vial  (Fraction 1).

                  7.3.3.8     Close the cartridge valve,  replace the collection
            vial, and  add 5  ml  of diethyl  ether/hexane (50/50,  v/v)  to the
            cartridge.  Slowly open the cartridge valve and collect the eluate
            into the collection vial (Fraction 2).

                  7.3.3.9     Adjust  the  final  volume  of  each  of  the  two
            fractions  to  a  known  volume   which   will   result   in  analyte
            concentrations appropriate for the project requirements (normally 1
            - 10 mL).  The fractions may be combined prior to final  adjustment
            of volume, if analyte fractionation is not required.  Table 4 shows
            compound  recoveries  for 1 g silica cartridges.   The cleaned up
            extracts are ready for analysis by Method 8081.


8.0   QUALITY CONTROL

      8.1   Refer to  Chapter One for specific  quality  control  procedures and
Method 3600 for cleanup procedures.

      8.2   A reagent blank (consisting of the elution solvents) must be passed
through the column or cartridge and checked for the compounds of interest, prior
to the use of this method.  This same performance check is required with each new
lot of adsorbent or  cartridges.  The  level  of interferences  must  be  below the
method detection limit before this method is performed on actual samples.

      8.3   The analyst must demonstrate that the compounds of interest are being
quantitatively recovered before applying this method to actual  samples.  See the
attached Tables for  acceptable recovery data.   For compounds that have not been
tested, recovery must be > 85%.

            8.3.1 Before  any  samples  are  processed   using  the  solid-phase
      extraction cartridges,  the efficiency of the cartridge must be  verified.
      A recovery check must be performed  using standards of the target analytes
      at  known concentration.  Only lots  of  cartridges  that  meet  the recovery
      criteria for the spiked compounds can be used to process the samples.

            8.3.2 A  check  should  also  be performed  on  each  individual  lot of
      cartridges and for every 300 cartridges of a particular lot.

      8.4   For  sample  extracts  that  are cleaned  up  using  this method,  the
associated quality control  samples should also be processed  through this cleanup
method.
                                   3630B  - 9                         Revision 2
                                                                September 1994

-------
9.0   METHOD PERFORMANCE

      9.1   Table 1  provides  performance  information on the  fractionation  of
phenolic derivatives using standard column chromatography.

      9.2   Table 2  provides  performance  information on the  fractionation  of
organochlorine pesticides/Aroclors using standard column chromatography.

      9.3   Table 3 shows recoveries of derivatized phenols obtained using 2 g
silica cartridges.

      9.4   Table  4  shows  recoveries  and  fractionation  of  organochlorine
pesticides obtained using 1  g  silica cartridges.


10.0  REFERENCES

      1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures
            for the Analysis of Pollutants Under the Clean Water Act; Final Rule
            and Interim Final  Rule and Proposed Rule," October 26, 1984.

      2.    U.S  EPA  "Evaluation  of  Sample  Extract Cleanup Using Solid-Phase
            Extraction Cartridges," Project Report, December 1989.
                                  3630B - 10                        Revision 2
                                                                September 1994

-------
                                    TABLE  1
                   SILICA GEL FRACTIONATION OF PFBB DERIVATIVES
                                        Percent Recovery by Fraction8

Parameter                          123
2-Chlorophenol
2-Nitrophenol
Phenol
2,4-Dimethylphenol
2,4-Dichlorophenol
2,4,6-Trichlorophenol
4-Chl oro -3 -methyl phenol
Pentachlorophenol
4-Nitrophenol
90

90
95
95
50 50
84
75 20

1
9
10
7
1

14

1

90






90
a  Eluant  composition:

    Fraction 1 - 15% toluene in hexane.
    Fraction 2 - 40% toluene in hexane.
    Fraction 3 - 75% toluene in hexane.
    Fraction 4 - 15% 2-propanol in toluene.

Data from Reference 1 (Method 604)
                                  3630B - 11                        Revision 2
                                                                September 1994

-------
    .a
    a
    GO
    z
 O
 ii
 D£ O
 O O
 O LU
    O
 oo
 LU -i O
 ct: i—i
 o GO
i o
I LU Z
 I— GO
 z a:
 LU O
 O —I
 ct: o
 LU O
 a. a:
    «£
 Q
 Z  
 O CO
 i— . O
 t— o.

 CO O
 i— i Z
 a: <
 I—
 GO GO
 »-H LU
 Q a
    »— 4
    O
    (J
 0)

 o
 o
 OJ
                    u
                    c
                    U
o

4->
u
                    o
                    o
                  Oinr-voi— voooencMcoioco^j-cvimcMr- r-. r—i ,—i o

                  COCMOOCOi—"Or-lOCMOCSIOOCncnoOOOlOin

                  ^- —••—«—oo »*  >—^oo ^-"a- >—•>-•>—••—10 CM •~'-—-—-r*. LO
                  rs-ootocoi—iooOCT»r^cnr^i—ivocMr^oocMcvii—ioocT>
                     CTl  00 CXJ t—i •—i  00    CO    OOr^OOCTi       OOOOO
                                                                          i—i



                  »^i—ivoif>i—tvoi— cO'-<'*oto^-vocyicj>ro 01 o o •—i

                  i—iCMroco^-m^-coLOLOioO'^-^i-'—i^-ir>CT»CM«s-«a-

                  CMr~-i—icMCT>r~.Lf>CMLOioio ^-T^ CM i—i ro •—i cr> •^-* vo •—i
                  ooocncriOCTicrivocnoocnmcrvooocnocnoocotTi
                     i—i       i—i                   00    •—i       I-H    00






                  omr^vo       CM •—i cvi    voco^i-cMLOCMOi^t^

                  00 CM  O O       O LO O    OCMOOCTtCTlLOOO
                  *—* i—I  r—I i—I       r-H •*—^r-H    i—* i—I •—t t—1 *—^^.^'i-H f-H i-H

                  r--ooinoo       oocot^    r^i—ivocMr^oor~>cM«a-
                     CDOOOO       00    00    00 t~- CO CT1         • 00 CO
                                                                   00


                  i-» i—i  vo ir>       r~- o •—i    om^»-«X)CT>crii~-cri'5i-

                  i—i CM  ro co       ^t LO in    lOO^-^-r-i^a-r^cricn

                  CMt-^i—ICM       LOCnkO    lO •	1— CVJ f—I CO •	CTl PO
                  ooocr>cr>       cncMcn    cnLocnococriLncnr^
                     •—i                             OO    i—i       i—i
                                        PO
                                                         en
                                                         ro
                                                                                                                      CM ^1-
                                                                                                                         cn
                                                                                                                      c cr>
                                                                                                                      O r-H

                                                                                                                      VI &-
                                                                                                                      •r- 0)
                                                                                                                      > J2
                                                                                                                      
                                                                                                                         a.
                                                                                                                         0)
                                                                                                                         oo
c
o
u
ro
U
C
o
o
                               r^ 10

                               00 I-H

                               oo »*•
                               •-i O
                               •—110
u
«o
                                        co

                                        10

                                        cr>
                                                         ro

                                                         in
                                        CM
                                        CM
                                        If)
                                         •

                                        in
                                                                    co
                                                                   UD
                                                                   00
                                                                          CM

                                                                          in
                                                                                                                      CM
                                                                                                  OQ
                                                                                                  O
                                                                                                  CO
                                                                                                  10
                                                                                                  CO
                               en r^.    *r
                               o en    •—i
                                              oo

                                              CM


                                              CM
                                              in
                                              oo
                                                                              .—i O

                                                                              10 m
                                                                           r-~ in
                                                                           00 en
                                                                             O i-H
                                                                             oo en
    GO
    LU
    O.
                  <_>
                  ac
                  co
                                       o
                                    I   CO
                                          ca
                    o
                    d.
                    o
                    o
                                         O)
                                      O)  £                            S-
                                     •r- T3                       CO     O
                                      X  S-                    CD <4-    i—
                                      O  O                    "O i—    -C
                                      O.1—              I-H    >> 3     O
                                      0) -C i—"           I-H    _C 00     >>IOO
                                         O                    0)        X    I-H IO
                                S-     S-     C           C    T3 C     O,-   O CM
                            c ^  Q     Q p—  lO           CO*_   ^~ COt*-   *" GJ r-H r-H
                            lEi—    r—  
-------
cu
c
cu

>,
_!_>





























CM CU
3
LU C
	 | .f—
CQ +->


t-H

C
o
o
(TJ
s-
1 1

cu
(TJ
X
CD

— J
E
0
CO

H- 1
c
o
•r—
+J
O
(TJ
s_
U_

C
o
•r—
£
C/)
O
Q.
E
0
O
•
-t-> cu
C T3
CD -i-
3 S-
i— O

It- !c
LU U

O TO i-
• c o
.— ( 03 <4_

•" - e
"H cu £
<= "5,,—
Oj_ -C O
i— cu
3 J2 i_
-g * S.
c c:
UJ .r- O1
T3 -0 ^
^ ^^
(U LU i— i
^
CD " **"
x -h -2
0 -O T3
Q_ C S—
27 LU o
. r-
0 h-T<->
sg-
(TJ J O
Q.^. C
CD «* -C
3: o
- cu
.JEo40
i» *"* C"
-o ^' m
< "^ s_
^ ^- o
s- rr
r^ Q "5
•^ - 4J
31 ^-^^
.'
j
«2 ^ **" T*
*3 C sL
^ 01 E O
^ 0 3 r-
-3 "O i— (_)
'TT C O o
o 1 1 i (.j ^_
**J ^
t »» i-
CD C g_~g
c_ fO
CT> "O D^
^ .^! 2
o .^^
t/1 ^" ^ t-
-«£^°
^-1 "+- U
c <— o
C E = S-
o 3 "> <:

4-> O C -
(TJ O T( CD
S- U- C
+-> 5_ r— CD
CD A. <" CL
U 0 TJ
C T3 X
o cy c o
O 3.LU H-
^1












^
i — i
C
o

+J
s-
4->
c:
CD
U
O
0
S-
O
**"
cu
oo
o
JC
00
ca
.c
cr>
f-

00
CU
•-

0
T-H

CU
(TJ

T3
CD
-r-
QL
00
oo
-i-J

3
0
(TJ
CD
.^Z
4->
OJ

C
o
• f—
(TJ
i_
4_3
£
CD
O
0
0
s-
o
u_
o
-o
^
(TJ
-o
c
_4
*V
00
CD
-C

cu
S-
(TJ

CU
00
cu

•^
c
CD
(TJ
CL
C
'""
I/)
S-
cu
3
C

• *s
00
o
[ '
c
•r-
g
S-
CD
4->
cu
-o
CD
s-
I t
•*- 4->
0 c
.. CD
r^> o
S- J_
cu cu
?; °-
o
0 to
CD
^ 00
CD "~
01 4->
c *i^"
flj *r~

(0 °-
5°
+-> "=
cu
CD i-
^_ ^
Q. >
cu o
*" CD
c s-
cu
> " "
•f— £
C71 O
oo +->
CD (TJ
3 •<-
r— >
(TJ CU
> -o
•D












t
CU
c
(O
X
cu
-C
_l
(NJ
c


cu
'o
00
00
•f—
•o
o

-o
c
(TJ
-Q

00
CD
O
C
0
o


.^
^
cu
1 ^
nj
O
-o
C
00
TJ

**
00
T3
S-
(t>
T3
m
oo
c~
£
2
-t-,
CD
C
•^
(T3
^— >
O
(TJ
4_)
(TJ
O
V
oo
00
O
S-
U
00
•o
C
3
O
CL
O
O .
fli
 O
(TJ as
•r- CD
S-
> -1-
-C -D
O CD
+J S-
(TJ m
-Q >
1 0
0 0
+-> CD
I S-
0 oo

fX5 g*
-Q 3
•si
O ro

A QJ
3 5
t_>
j^J
5 o
^^ »r~
-o +j
^ ^3
£^=

c o
CD S
CD +-1
-Q
C
oo •"-
(TJ
CD
4_> ^
<-" 0

CVJ ^t
   o»
 c 0%
 O i-H

 00  S-
•i-  CU
 > J3
 CU  E
OL  CU
   4->
    Q.
    CD
   CO
oo
o
CO
vo
CO

-------
                                      TABLE  3
                  PERCENT RECOVERIES AND ELUTION PATTERNS FOR 18
                        PHENOLS FROM 2 g SILICA CARTRIDGES3
                                                    Fraction 2
                                                Average     Percent
      Compound                                  Recovery      RSD
Phenol
2-Methyl phenol
3 -Methyl phenol
4-Methyl phenol
2, 4-Dimethyl phenol
2-Chlorophenol
2,6-Dichlorophenol
4-Chl oro-3-methyl phenol
2,4-Dichlorophenol
2,4,6-Trichlorophenol
2 , 3 , 6-Tri chl orophenol
2,4,5-Trichlorophenol
2 , 3 , 5-Tri chl orophenol
2,3,5,6-Tetrachlorophenol
2,3,4,6-Tetrachlorophenol
2, 3, 4-Tri chl orophenol
2,3,4, 5-Tetrachl orophenol
Pentachl orophenol
74.1
84.8
86.4
82.7
91.8
88.5
90.4
94.4
94.5
97.8
95.6
92.3
92.3
97.5
97.0
72.3
95.1
96.2
5.2
5.2
4.4
5.0
5.6
5.0
4.4
7.1
7.0
6.6
7.1
8.2
8.2
5.3
6.1
8.7
6.8
8.8
a     Silica cartridges  (Supelco,  Inc.)  were  used;  each cartridge was conditioned
      with 4 mL of hexane prior to use.   Each  experiment  was  performed  in duplicate
      at three spiking concentrations (0.05 ptg,  0.2  jug,  and 0.4  y,q per  compound  per
      cartridge).   Fraction  1  was  eluted with  5  mL  hexane  and  was discarded.
      Fraction 2 was eluted with 5 mL toluene/hexane (25/75,  v/v).

Data from Reference 2
                                    3630B  -  14                           Revision  2
                                                                     September 1994

-------
                                      TABLE 4
           PERCENT RECOVERIES AND ELUTION PATTERNS FOR 17 ORGANOCHLORINE
                PESTICIDES AND AROCLORS FROM 1 g SILICA CARTRIDGES3
Compound
     Fraction 1
Average     Percent
Recovery      RSD
     Fraction 2
Average     Percent
Recovery      RSD
alpha-BHC
gamma -BHC
beta-BHC
Heptachlor
delta-BHC
Aldrin
Heptachlor epoxide
Endosulfan I
4,4'-DDE
Dieldrin
Endrin
4,4'-DDD
Endosulfan II
4,4'-DDT
Endrin aldehyde
Endosulfan sulfate
4,4'-Methoxychlor
Aroclor 1016
Aroclor 1221
Aroclor 1232
Aroclor 1242
Aroclor 1248
Aroclor 1254
Aroclor 1264
0
0
0
97.3 1.3
0
95.9 1.0
0
0
99.9 1.7
0
0
10.7 41
0
94.1 2.0
0
0
0
124
93.5
118
116
114
108
112
98.7
94.8
94.3
0
90.8
0
97.9
102
0
92.3
117
92.4
96.0
0
59.7
97.8
98.0







2.3
1.9
3.0

2.5

2.1
2.3

2.0
2.6
3.3
2.2

2.6
2.1
2.4







a     Silica cartridges  (Supelco,  Inc.  lot SP0161) were  used;  each cartridge was
      conditioned with 4 mL hexane  prior to use.  The organochlorine pesticides were
      tested separately  from  PCBs.   Each  organochlorine pesticides experiment was
      performed in duplicate, at three spiking concentrations (0.2 ;ug, 1.0 /xg, and
      2.0 fj,g  per compound  per cartridge).   Fraction 1  was  eluted with  5  mL of
      hexane,  Fraction 2 with 5 mL  of diethyl  ether/hexane (50/50,  v/v).   PCBs were
      spiked at 10 ng per cartridge and were eluted with 3 mL of hexane.   The  values
      given for PCBs are the percent recoveries for a single determination.

Data from Reference 2
                                    3630B - 15
                                          Revision 2
                                      September 1994

-------
   METHOD 3630B
SILICA GEL CLEANUP
oc
PCBs
y/Com
> 10-30 mg 	 (
I ^\,nt.r
7.2
Colurr
Standard
n Cleanup.
Pesticide
ft. Phenols
7.I.N.
emrationN.
"' J
ferences./
( StSrt )
/ 7.1.1 N.
, .. ,( An«|vte \
N. Type. /

j/T\
VJ
I
\_/ 7.3 Cartridge
Cleanup.
y/Analyte N^ 	
I C Type. .? 	 1
Denvatized >^ / OC Pesticides
Phenols ^*S ft. PCBs
1 +
7.2.2.1 Do PFBB
derivatization on
sample extract
(8040).
4
7.2.2.2 Place
activated silica gel
in chromatographic
column; add
anhydroua NajSO*.
4
7.2.2.3 Preelute
column with hexane;
pipet hexane
solution onto column;
alute.
1
7.2.2.4 Elute column
with specified
solvents.
4
Analyze
by GC
(Method I
V. 8040). }
7.2.3.1 Deactivate
silica gel, prepare
column.
1
7.2.3.2 Elute the
GC column
with hexane.
V
7.2.3.3 Transfer
extract onto column
snd elute with
specified solvents.
^
7.3.4 Exchange the
elution solvent
to hexane (Section
7.1.3).
|
Analyze
by GC
I Method
N. 8081. >


i
7.3.1 Cartridge
Set-up ft.
Conditioning.
X'Analyte \^
|—C TVP.. >— |
Derivatized >. / OC Peaticides
Phenols N./ ft PCBs
1 1
7.3.2.1 Do PFBB
derivatization on
sample extract
(8040).
4
7.3.2.3 ft 7.3.2.4
Transfer extract
to cartridge.
\r
7.3.2.6 ft 7.3.2.7
Rinse cartridge
with hexane ft
discard.
1
7.3.2.8 Elute
cartridge with
toluene/he xane.
^f
Analyze by
GC Method
8040 or
GC/MS
\ Method /
\8270./
7.3.3.1 Exchange
solvent to
hexane.
^ r
7.3.3.3 & 7.3.3.4
Transfer extract
to cartridge.
4
7.3.3.6 & 7.3.3.7
Elute cartridge
with hexane as
Fraction 1.
1
7.3.3.8 Elute
cartridge with
ether/hexane as
Fraction II.
V
Analyze
each fraction
by GC
Method
\*™./
    3630B  -  16
    Revision I
September 1994

-------
 METHOD  3630B
  (continued)
   7.2 Standard
  Column Cleanup.
 7.2.1.1 Exchange
 extract solvent to
cyclohexane during
  K-D procedure.
 7.2.1.2 Prepare
 slurry activated
silica gel, prepare
     column.
  7.2.1.3 Preelute
   column with
 pentane, transfer
extract onto column
  and elute with
     pentane.
   7.2.1.4 Elute
   column with
 CH2CI2 /pentane;
    concentrate
 collected fraction;
   adjust volume.
     Analyze
   by GC Method
     8100 or
      GC/MS
     Method
      8270.
    3630B  -  17
     Revision  2
September  1994

-------
                                 METHOD 3640A

                            GEL-PERMEATION CLEANUP
1.0   SCOPE AND APPLICATION

      1.1   Gel-permeation  chromatography  (GPC)  is  a size  exclusion  cleanup
procedure  using  organic solvents  and  hydrophobic gels  in the  separation  of
synthetic macromolecules (1).  The packing gel is  porous and is characterized by
the range or uniformity (exclusion range) of that pore size.   In the choice of
gels, the exclusion range must be larger than the molecular size of the molecules
to be separated (2).  A  cross-linked divinylbenzene-styrene copolymer (SX-3 Bio
Beads or equivalent)  is specified for this method.

      1.2   General  cleanup application - GPC is recommended for the elimination
from the sample of lipids,  polymers, copolymers,  proteins, natural  resins and
polymers, cellular components, viruses, steroids, and dispersed high-molecular-
weight compounds (2).  GPC is appropriate  for both polar and non-polar analytes,
therefore, it  can be  effectively used  to cleanup extracts containing  a broad
range of analytes.

      1.3   Specific application -  This method  includes guidance  for cleanup of
sample extracts containing the following analytes  from the RCRA Appendix VIII and
Appendix IX lists:
      Compound Name                                            CAS No.'
      Acenaphthene                                             83-32-9
      Acenaphthylene                                          208-96-8
      Acetophenone                                             98-86-2
      2-Acetylaminofluorene                                    53-96-3
      Aldrin                                                  309-00-2
      4-Aminobiphenyl                                          92-67-1
      Aniline                                                  62-53-3
      Anthracene                                              120-12-7
      Benomyl                                                17804-35-2
      Benzenethiol                                            108-98-5
      Benzidine                                                92-87-5
      Benz(a)anthracene                                        56-55-3
      Benzo(b)fluoranthene                                    205-99-2
      Benzo(a)pyrene                                           50-32-8
      Benzo(ghi)perylene                                      191-24-2
      Benzo(k)fluoranthene                                    207-08-9
      Benzoic acid                                             65-85-0
      Benzotrichloride                                         98-07-7
      Benzyl alcohol                                          100-51-6
      Benzyl chloride                                         100-44-7
      alpha-BHC                                               319-84-6
      beta-BHC                                                319-85-7
                                  3640A - 1                         Revision 1
                                                                September 1994

-------
Compound Name
gamma-BHC
delta-BHC
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
2-sec-butyl -4,6-dinitrophenol (Dinoseb)
Carbazole
Carbendazim
alpha-Chlordane
gamma-Chlordane
4-Chloro-3-methyl phenol
4-Chloroaniline
Chi orobenzi late
Bi s (2-chl oroethoxy )methane
Bis(2-chloroethyl) ether
Bis (2-chl oroisopropyl) ether
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenol
3-Chlorophenol
4-Chlorophenyl phenyl ether
3-Chloropropionitrile
Chrysene
2-Cresol
3-Cresol
4-Cresol
Cyclophosphamide
ODD
DDE
DDT
Di-n-butyl phthalate
Dial late
Dibenzo(a,e)pyrene
Dibenzo(a,i)pyrene
Dibenz(a,j)acridine
Dibenz (a, h) anthracene
Dibenzofuran
Dibenzothiophene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
trans- l,4-Dichloro-2-butene
cis-l,4-Dichloro-2-butene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3, 3 '-Dichl orobenzi dine
2,6-Dichlorophenol
2,4-Dichlorophenoxyacetic acid (2,4-D)
2,4-Dichlorophenol
CAS No.a
58-89-9
319-86-8
101-55-3
85-68-7
88-85-7
86-74-8
10605-21-7
5103-71-9
5566-34-7
59-50-7
106-47-8
510-15-6
111-91-1
111-44-4
108-60-1
91-58-7
95-57-8
106-48-9
108-43-0
7005-72-3
542-76-7
218-01-9
95-48-7
108-39-4
106-44-5
50-18-0
72-54-8
72-55-9
50-29-3
84-74-2
2303-16-4
192-65-4
189-55-9
224-42-0
53-70-3
132-64-9
132-65-0
96-12-8
106-93-4
110-57-6
1476-11-5
95-50-1
106-46-7
541-73-1
91-94-1
87-65-0
94-75-7
120-83-2
3640A - 2
    Revision 1
September 1994

-------
Compound Name
2,4-Dichlorotoluene
l,3-Dichloro-2-propanol
Dieldrin
Diethyl phthalate
Dimethoate
Dimethyl phthalate
p-Di methyl ami noazobenzene
7, 12-Dimethyl -benz(a)anthracene
2,4-Dimethylphenol
3,3-Dimethylbenzidine
4,6-Dinitro-o-cresol
1,3-Dinitrobenzene
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diphenylamine
Diphenyl ether
1 , 2 -Di phenyl hydrazi ne
Disulfoton
Endosulfan sulfate
Endosulfan I
Endosulfan II
Endrin
Endrin aldehyde
Endrin ketone
Ethyl methane sulfonate
Ethyl methacrylate
Bis(2-ethylhexyl) phthalate
Famphur
Fluorene
Fluoranthene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachl orobutadi ene
Hexachl orocycl opentadi ene
Hexachloroethane
Hexachl oropropene
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
cis-Isosafrole
trans-Isosafrole
Kepone
Malononitrile
Merphos
Methoxychlor
3-Methylcholanthrene
CAS No.a
95-73-8
96-23-1
60-57-1
84-66-2
60-51-5
131-11-3
60-11-7
57-97-6
105-67-9
119-93-7
534-52-1
99-65-0
51-28-5
121-14-2
606-20-2
122-39-4
101-84-8
122-66-7
298-04-4
1031-07-8
959-98-8
33213-65-9
72-20-8
7421-93-4
53494-70-5
62-50-0
97-63-2
117-81-7
52-85-7
86-73-7
206-44-0
76-44-8
1024-57-3
118-74-1
87-68-3
77-47-4
67-72-1
1888-71-7
193-39-5
465-73-6
78-59-1
17627-76-8
4043-71-4
143-50-0
109-77-3
150-50-5
72-43-5
56-49-5
3640A - 3
    Revision 1
September 1994

-------
Compound Name
2-Methyl naphthal ene
Methyl parathion
4,4'-Methylene-bis(2-chloroaniline)
Naphthalene
1,4-Naphthoquinone
2-Naphthylamine
1-Naphthylamine
5-Nitro-o-toluidine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
2-Nitrophenol
4-Nitrophenol
N-Nitrosodi-n-butylamine
N-Nitrosodiethanolamine
N-Nitrosodi ethyl ami ne
N-Nitrosodi methyl ami ne
N-Ni trosodi phenyl ami ne
N-Nitrosodi-n-propylamine
N-Ni trosomethyl ethyl ami ne
N-Ni trosomorphol ine
N-Nitrosopiperidine
N-Nitrosopyrolidine
Di-n-octyl phthalate
Parathion
Pentachl orobenzene
Pentachl oroethane
Pentachloronitrobenzene (PCNB)
Pentachl orophenol
Phenacetin
Phenanthrene
Phenol
1 , 2- Phenyl enedi ami ne
Phorate
2-Picoline
Pronamide
Pyrene
Resorcinol
Safrole
1,2,4, 5 -Tetrachl orobenzene
2,3,5,6-Tetrachloronitrobenzene
2,3,5, 6 -Tetrachl orophenol
2,3,4,6-Tetrachlorophenol
Tetraethyl dithiopyrophosphate (Sulfotep)
Thiosemicarbazide
2-Toluidine
4-Toluidine
CAS No.a
91-57-6
298-00-0
101-14-4
91-20-3
130-15-4
91-59-8
134-32-7
99-55-8
88-74-4
99-09-2
100-01-6
98-95-3
79-46-9
100-02-7
924-16-3
1116-54-7
55-18-5
62-75-9
86-30-6
621-64-7
10595-95-6
59-89-2
100-75-4
930-55-2
117-84-0
56-38-2
608-93-5
76-01-7
82-68-8
87-86-5
62-44-2
85-01-8
108-95-2
95-54-5
298-02-2
109-06-8
23950-58-5
129-00-0
108-46-3
94-59-7
95-94-3
117-18-0
935-95-5
58-90-2
3689-24-5
79-19-6
106-49-0
95-53-4
3640A - 4
    Revision 1
September 1994

-------
      Compound Name                                             CAS No.'
      Thiourea,  l-(o-chlorophenyl)                            5344-82-1
      Toluene-2,4-diamine                                       95-80-7
      1,2,3-Trichlorobenzene                                    87-61-6
      1,2,4-Trichlorobenzene                                   120-82-1
      2,4,6-Trichlorophenol                                     88-06-2
      2,4,5-Trichlorophenol                                     95-95-4
      2,4,5-Trichlorophenoxyacetic acid  (2,4,5-T)               93-76-5
      2,4,5-Trichlorophenoxypropionic acid  (2,4,5-TP)           93-72-1
      Warfarin                                                  81-81-2
      a  Chemical Abstract Services Registry Number.

      Table 1 presents average percent recovery and percent RSD data for these
analytes,  as  well  as the  retention volumes of  each  analyte on  a  single GPC
system.   Retention volumes  vary  from  column  to column.   Figure  1 provides
additional information on  retention volumes for certain classes of compounds.
The data for the semivolatiles were determined  by  GC/MS, whereas, the pesticide
data were  determined  by  GC/ECD or  GC/FPD.   Compounds  not amenable  to GC were
determined by HPLC.   Other analytes may  also  be  appropriate  for this cleanup
technique, however, recovery through the  GPC should be >70%.

      1.4   Normally, this method  is most efficient for removing high boiling
materials that condense in the injection  port area of a gas chromatograph (GC)
or  the  front  of  the  GC  column.    This  residue will  ultimately  reduce  the
chromatographic separation efficiency or  column capacity because of adsorption
of the target analytes on  the  active sites.  Pentachlorophenol  is especially
susceptible to this problem.  GPC,  operating on  the principal of size exclusion,
will not usually remove interference peaks that  appear in the chromatogram since
the molecular size  of these compounds is relative similar to the target analytes.
Separation cleanup techniques, based on other molecular characteristics (i.e.,
polarity), must be used to eliminate this type  of interference.


2.0   SUMMARY OF METHOD

      2.1   The  column   is  packed with  the  required amount  of  preswelled
absorbent, and is  flushed  with solvent for  an  extended period.   The column is
calibrated and then loaded  with the sample extract to  be cleaned up.  Elution is
effected with a suitable solvent(s) and the product  is then concentrated.


3.0   INTERFERENCES

      3.1   A reagent  blank should be analyzed for the compound of interest prior
to the  use of  this method.   The  level  of interferences  must  be  below  the
estimated quantitation limits  (EQLs)  of  the analytes  of  interest  before  this
method is performed on actual samples.


                                   3640A - 5                        Revision 1
                                                                September 1994

-------
      3.2   More extensive procedures than  those outlined  in this method may be
necessary for reagent purification.


4.0   APPARATUS

      4.1   Gel-permeation chromatography  system  -  GPC Autoprep Model  1002  A
or B, or equivalent,  Analytical  Biochemical  Laboratories, Inc.   Systems  that
perform  very  satisfactorily  have  also  been assembled  from  the  following
components -  an HPLC pump, an auto sampler or a valving system with sample loops,
and a fraction collector.  All systems, whether automated or manual, must meet
the calibration requirements of Sec. 7.2.2.

            4.1.1 Chromatographic column -  700 mm x 25 mm ID glass column.  Flow
      is upward.   (Optional)   To  simplify switching from the UV detector during
      calibration to the GPC  collection device during extract cleanup, attach a
      double  3-way  valve  (Rheodyne Type  50  Teflon  Rotary  Valve  #10-262  or
      equivalent) so that the column exit flow can be shunted either to the UV
      flow-through cell or to the GPC collection  device.

            4.1.2 Guard column -  (Optional)  5 cm, with appropriate fittings to
      connect  to  the  inlet side  of the analytical  column (Supelco  5-8319  or
      equivalent).

            4.1.3 Bio Beads (S-X3)  - 200-400 mesh, 70 g (Bio-Rad Laboratories,
      Richmond, CA,  Catalog 152-2750 or equivalent).   An additional 5 g of Bio
      Beads are required if the optional guard column  is employed.  The quality
      of Bio Beads may vary from lot to lot because of excessive fines in some
      lots.   The  UV chromatogram  of  the  Calibration solution  should  be  very
      similar  to  that  in Figure  2,  and  the  backpressure should be  within  6-
      10 psi.  Also,  the gel  swell  ratio in methylene chloride should be in the
      range of 4.4 - 4.8 mL/g.  In addition  to  fines having a detrimental effect
      on chromatography,  they can  also pass through  the column  screens  and
      damage the valve.

            4.1.4 Ultraviolet detector - Fixed wavelength  (254 nm) with a semi-
      prep flow-through cell.

            4.1.5 Strip chart recorder, recording  integrator or laboratory data
      system.

            4.1.6 Syringe - 10 mL with Luerlok fitting.

            4.1.7 Syringe  filter assembly, disposable  -  Bio-Rad  "Prep  Disc"
      sample  filter  assembly  #343-0005, 25 mm,  and  5 micron filter discs  or
      equivalent.   Check  each   batch  for contaminants.    Rinse  each  filter
      assembly (prior to use) with methylene chloride if necessary.

      4.2   Analytical balance - 0.0001 g.

      4.3   Volumetric flasks, Class A - 10 mL to 1000 mL

      4.4   Graduated cylinders


                                   3640A -  6                         Revision 1
                                                                September 1994

-------
5.0  REAGENTS

      5.1   Methylene chloride, CH2C12.   Pesticide  quality  or  equivalent.

            5.1.1 Some  brands  of methylene chloride may  contain unacceptably
      high  levels  of acid (HC1).   Check the  pH by shaking equal  portions of
      methylene chloride and water, then check the pH of the water layer.

                  5.1.1.1     If the pH  of  the  water  layer is < 5,  filter the
            entire  supply  of solvent through  a 2  in.  x  15  in.  glass column
            containing  activated  basic  alumina.    This  column  should  be
            sufficient  for  processing  approximately 20-30 liters  of solvent.
            Alternatively, find a different supply of methylene chloride.

      5.2   Cyclohexane, C6H12.   Pesticide quality  or equivalent.

      5.3   n-Butyl chloride, CHaC^C^CHjCl.  Pesticide quality or equivalent.

      5.4   GPC  Calibration  Solution.    Prepare  a  calibration solution  in
methylene chloride containing the following analytes (in elution order):

      Compound                              mg/L
      corn oil                            25,000
      bis(2-ethylhexyl) phthalate          1,000
      methoxychlor                           200
      perylene                                20
      sulfur                                  80

      NOTE: Sulfur  is  not  very  soluble  in methylene chloride,  however,  it is
            soluble in warm corn oil.  Therefore, one approach is to weigh out
            the corn oil, warm it and transfer the weighed amount  of sulfur into
            the warm corn  oil.  Mix it and then transfer  into a volumetric flask
            with methylene chloride, along with the other calibration compounds.

      Store the calibration solution in an amber glass bottle with a Teflon lined
screw-cap at 4°C, and protect from light.   (Refrigeration may cause the corn oil
to precipitate.  Before use,  allow the calibration solution  to  stand  at room
temperature until the  corn  oil  dissolves.)   Replace  the  calibration standard
solution every 6 months, or more frequently if necessary.

      5.5   Corn Oil Spike for Gravimetric Screen.  Prepare a  solution  of corn
oil in methylene chloride (5 g/100 ml).


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory material  to this chapter, Organic Analytes, Sec.
4.1.
                                  3640A  - 7                         Revision 1
                                                                September 1994

-------
7.0  PROCEDURE

      7.1   It  is  very important  to  have consistent  laboratory  temperatures
during an entire GPC run,  which could  be  24 hours or more.  If temperatures are
not consistent,  retention times will  shift,  and  the  dump and collect  times
determined by the calibration standard will no longer be appropriate.  The ideal
laboratory temperature to prevent outgassing  of  the methylene chloride is 72°F.

      7.2   GPC Setup and Calibration

            7.2.1 Column Preparation

                  7.2.1.1     Weigh out 70 g of Bio Beads  (SX-3).  Transfer them
            to a quart  bottle  with a  Teflon  lined cap or  a  500 ml separatory
            funnel  with a large bore stopcock, and add approximately 300 ml of
            methylene chloride.  Swirl the container  to  ensure  the wetting of
            all  beads.   Allow the  beads  to  swell  for a minimum  of  2  hours.
            Maintain  enough  solvent to  sufficiently  cover  the  beads at  all
            times.   If a guard column is  to be used, repeat the above with 5 g
            of  Bio  Beads  in  a  125  ml  bottle  or  a  beaker, using  25 ml  of
            methylene chloride.

                  7.2.1.2     Turn  the column  upside down  from  its  normal
            position,  and remove  the  inlet  bed   support  plunger   (the  inlet
            plunger is longer  than the outlet plunger).   Position  and tighten
            the outlet bed support plunger as near the end as possible,  but no
            closer than 5 cm (measured from the  gel packing to  the  collar).

                  7.2.1.3     Raise the  end  of the outlet  tube  to  keep  the
            solvent in the GPC column, or close the column outlet  stopcock if
            one is attached.   Place a small  amount of solvent in the column to
            minimize the formation of air bubbles  at the  base of poured column
            packing.

                  7.2.1.4     Swirl the bead/solvent slurry to get a homogeneous
            mixture and,  if  the wetting  was  done  in  a  quart  bottle,  quickly
            transfer  it   to  a  500 ml  separatory  funnel  with a  large  bore
            stopcock.   Drain  the excess  methylene  chloride  directly  into the
            waste beaker,  and then  start  draining the slurry into the column by
            placing the separatory  funnel  tip against the column  wall.   This
            will help to minimize bubble  formation.  Swirl occasionally to keep
            the slurry homogeneous.  Drain enough to fill  the  column.  Place the
            tubing from the column  outlet into a waste  beaker below the column,
            open the  stopcock  (if attached)   and  allow the excess  solvent to
            drain.   Raise the tube  to  stop the flow and close the stopcock when
            the top of  the  gel begins to look dry.   Add additional  methylene
            chloride to just rewet the gel.

                  7.2.1.5     Wipe  any  remaining   beads  and  solvent  from  the
            inner walls  of the  top  of the  column with  a  laboratory tissue.
            Loosen  the seal  slightly on the  other  plunger  assembly  (long
            plunger) and  insert  it  into  the  column.   Make the  seal just tight
                                   3640A  -  8                         Revision 1
                                                                September 1994

-------
enough  so  that  any beads  on the  glass  surface  will  be pushed
forward, but loose enough so that the plunger can  be pushed  forward.

      CAUTION:    Do not tighten the seal if beads are between the
                  seal  and  the  glass  surface  because this  can
                  damage the seal and cause leakage.

      7.2.1.6     Compress the column  as  much  as possible without
applying excessive  force.  Loosen  the  seal  and gradually pull out
the plunger.  Rinse and wipe  off  the  plunger.   Slurry any remaining
beads and transfer  them into  the column.   Repeat Sec.  7.2.1.5 and
reinsert the plunger.  If  the  plunger cannot be inserted and pushed
in  without  allowing  beads to  escape  around  the  seal,  continue
compression of the beads without tightening the  seal,  and loosen and
remove the plunger  as  described.   Repeat  this  procedure until the
plunger is successfully inserted.

      7.2.1.7     Push the  plunger  until  it  meets the  gel,  then
compress the column bed about four centimeters.

      7.2.1.8     Pack the optional  5 cm column with approximately
5 g  of preswelled   beads  (different  guard columns   may  require
different amounts).  Connect  the guard column  to the inlet of the
analytical  column.

      7.2.1.9     Connect the column  inlet  to the solvent reservoir
(reservoir should be placed higher than the top of the column) and
place  the  column outlet  tube in  a  waste  container.    Placing  a
restrictor in the outlet tube  will  force air out of the column more
quickly.   A  restrictor  can  be  made  from a  piece of  capillary
stainless steel  tubing of 1/16"  OD x  10/1000"  ID x 2".   Pump
methylene chloride through the column at  a  rate of 5 mL/min for one
hour.

      7.2.1.10    After washing  the  column  for at  least one hour,
connect the  column outlet tube, without  the restrictor,  to the inlet
side of the UV detector.   Connect  the  system outlet to the outlet
side  of the  UV  detector.   A restrictor  (same  size  as  in  Sec.
7.2.1.9) in  the outlet  tube from the UV  detector will prevent bubble
formation which  causes  a noisy UV baseline.   The restrictor will not
effect  flow  rate.    After  pumping methylene chloride  through the
column for  an  additional  1-2  hours,  adjust the  inlet  bed  support
plunger until  approximately 6-10 psi backpressure is  achieved.  Push
the plunger  in to  increase pressure or slowly pull outward to reduce
pressure.

      7.2.1.11    When the GPC column is not to be used for several
days, connect the column outlet line  to the column inlet to prevent
column drying  and/or channeling.  If channeling  occurs,  the gel must
be removed  from  the column,  reswelled, and repoured  as  described
above.   If  drying   occurs,  methylene  chloride  should be pumped
through the column  until the  observed  column pressure  is constant
and the column appears  wet.  Always recalibrate after column drying
has occurred to verify retention volumes  have not changed.

                       3640A  -  9                         Revision 1
                                                    September 1994

-------
7.2.2 Calibration of the GPC Column

      7.2.2.1     Using a 10 ml  syringe,  load  sample  loop #1 with
calibration solution (Sec. 5.6).  With the ABC automated system, the
5 ml  sample loop requires  a  minimum of  8 ml of  the calibration
solution.  Use a firm, continuous pressure to push the sample onto
the loop.  Switch the valve so that GPC flow is  through the UV flow-
through cell.

      7.2.2.2     Inject the calibration  solution  and obtain a UV
trace  showing  a discrete  peak  for  each  component.   Adjust  the
detector and/or recorder sensitivity to produce a UV trace similar
to Figure  2 that meets  the  following requirements.   Differences
between manufacturers' cell  volumes and detector sensitivities may
require a dilution  of  the calibration  solution to achieve similar
results.  An analytical  flow-through  detector  cell  will  require a
much  less   concentrated  solution  than  the semi-prep  cell,  and
therefore the analytical cell  is not acceptable for use.

      7.2.2.3     Following are  criteria  for  evaluating the  UV
chromatogram for column condition.

            7.2.2.3.1   Peaks   must  be   observed,   and  should  be
      symmetrical,  for all  compounds in the calibration solution.

            7.2.2.3.2   Corn oil and  phthalate  peaks  must exhibit
      >85% resolution.

            7.2.2.3.3   Phthalate  and   methoxychlor   peaks   must
      exhibit >85% resolution.

            7.2.2.3.4   Methoxychlor and perylene peaks must exhibit
      >85% resolution.

            7.2.2.3.5   Perylene  and  sulfur   peaks  must  not  be
      saturated and must exhibit >90% baseline resolution.

            7.2.2.3.6   Nitroaromatic  compounds   are  particularly
      prone to adsorption.  For  example,  4-nitrophenol  recoveries
      may be low due  to a portion of the analyte being  discarded
      after  the  end of  the  collection  time.   Columns  should be
      tested with the  semivolatiles  matrix spiking solution.  GPC
      elution should continue until  after perylene has eluted, or
      long enough to recover at least 85% of the analytes,  whichever
      time is longer.

      7.2.2.4     Calibration   for   Semivolatiles  -  Using  the
information from the  UV trace,   establish  appropriate collect and
dump  time  periods   to  ensure  collection  of all  target  analytes.
Initiate   column  eluate  collection   just  before   elution   of
bis(2-ethylhexyl) phthalate and  after the elution of the corn oil.
Stop  eluate collection  shortly  after  the  elution  of  perylene.
Collection  should be  stopped  before sulfur elutes.   Use  a  "wash"
time of 10 minutes  after the elution  of sulfur.  Each  laboratory is

                      3640A -  10                        Revision 1
                                                    September 1994

-------
required to establish its specific time sequences.  See Figure 2 for
general guidance on retention time.   Figure 1 illustrates retention
volumes for different classes of compounds.

      7.2.2.5     Calibration for Organochlorine Pesticides/PCBs -
Determine  the  elution  times  for   the  phthalate,  methoxychlor,
perylene, and sulfur.  Choose a  dump  time which  removes >85% of the
phthalate, but  collects  >95% of the  methoxychlor.  Stop collection
after the elution of perylene, but before sulfur elutes.

      7.2.2.6     Verify the flow rate by collecting column eluate
for 10 minutes in  a graduated cylinder and measure the volume, which
should be 45-55 ml (4.5-5.5 mL/min).  If the flow rate  is outside of
this range,  corrective  action must  be taken,  as described above.
Once the flow rate is within the range of 4.5-5.5 mL/min, record the
column pressure (should be  6-10  psi)  and room temperature.  Changes
in  pressure,  solvent  flow  rate,  and  temperature  conditions  can
affect analyte retention times,  and must be monitored.  If the flow
rate and/or column pressure do not fall within the above ranges, a
new column should be  prepared.   A UV trace  that does  not meet the
criteria  in  Sec.  7.2.2.3  would also  indicate  that a  new column
should be prepared.  It may  be necessary to obtain  a new lot of Bio
Beads if the column fails all the criteria.

      7.2.2.7     Reinject    the   calibration    solution    after
appropriate collect and dump cycles have been set,  and the solvent
flow and column pressure have been established.

            7.2.2.7.1   Measure and record the volume of collected
      GPC eluate in a graduated  cylinder.  The volume of GPC eluate
      collected for  each  sample extract  processed may  be  used to
      indicate problems with the system during sample processing.

            7.2.2.7.2   The  retention  times  for bis(2-ethylhexyl)
      phthalate and  perylene  must  not vary more than  +5%  between
      calibrations.   If  the retention  time shift  is  >5%,  take
      corrective action.  Excessive retention time shifts are caused
      by:

                  7.2.2.7.2.1 Poor 1 aboratory temperature control or
            system leaks.

                  7.2.2.7.2.2 An unstabilized column that requires
            pumping methylene chloride through it for several  more
            hours or overnight.

                  7.2.2.7.2.3 Excessive  laboratory  temperatures,
            causing outgassing of the methylene chloride.

      7.2.2.8     Analyze a GPC  blank by loading 5  ml  of methylene
chloride  into  the GPC.  Concentrate the  methylene chloride  that
passes through the system during  the  collect cycle using a Kuderna-
Danish  (KD)   evaporator.    Analyze  the  concentrate  by  whatever
detectors will be  used for the analysis of future samples.  Exchange

                      3640A - 11                        Revision 1
                                                    September 1994

-------
      the  solvent,  if necessary.   If the  blank exceeds  the  estimated
      quantitation  limit of  the  analytes,   pump  additional  methylene
      chloride through  the  system for 1-2  hours.   Analyze  another GPC
      blank  to  ensure  the  system  is  sufficiently clean.   Repeat the
      methylene chloride pumping, if necessary.

7.3   Extract Preparation

      7.3.1 Adjust the extract  volume  to 10.0 ml.  The  solvent extract
must  be  primarily  methylene  chloride.    All   other  solvents,  e.g.
1:1 methylene chloride/acetone,  must be concentrated to 1  ml (or  as low as
possible if  a precipitate  forms)  and  diluted to 10.0  ml with  methylene
chloride.  Thoroughly mix the extract  before proceeding.

      7.3.2 Filter the extract through a 5 micron  filter disc by attaching
a syringe filter assembly containing the filter disc to a 10 ml syringe.
Draw the sample  extract  through the filter assembly and  into  the 10 ml
syringe.  Disconnect  the filter assembly before  transferring the sample
extract  into a small  glass  container,  e.g.  a 15 ml culture  tube with a
Teflon lined screw cap.   Alternatively, draw the extract  into the syringe
without  the  filter  assembly.   Attach  the filter assembly  and  force the
extract through the filter  and  into the  glass  container.   The  latter is
the preferred technique  for viscous  extracts or extracts  with  a lot of
solids.  Particulate  larger than  5 microns may  scratch the valve, which
may result in a  system leak  and  cross-contamination of  sample extracts in
the sample loops.  Repair of the damaged valve is quite expensive.

      NOTE: Viscosity of a  sample extract should not exceed the viscosity
            of  1:1  water/glycerol.   Dilute  samples  that exceed  this
            viscosity.

7 A   Screening the Extract

      7.4.1 Screen  the extract  to determine  the  weight of  dissolved
residue  by  evaporating  a  100  juL  aliquot to  dryness  and weighing the
residue.  The weight of dissolved residue loaded on the GPC column cannot
exceed 0.500 g.   Residues  exceeding 0.500 g will very likely  result in
incomplete extract cleanup  and  contamination  of  the  GPC  switching valve
(which results in cross-contamination  of sample extracts).

            7.4.1.1     Transfer  100  /zL  of  the  filtered extract  from
      Sec.  7.3.2 to a tared aluminum weighing dish.

            7.4.1.2     A suggested evaporation technique is to  use a heat
      lamp.   Set  up  a  250 watt  heat   lamp  in  a  hood  so  that  it  is
      8 + 0.5 cm from  a  surface covered  with a clean  sheet  of aluminum
      foil.  Surface temperature should  be 80-100°C (check temperature by
      placing a thermometer on  the foil  and  under  the  lamp).   Place the
      weighing dish under the lamp using tongs.   Allow it to stay under
      the lamp for  1  min.   Transfer the  weighing dish to an analytical
      balance or a micro balance and weigh to  the nearest 0.1 mg.  If the
      residue weight is less than 10 mg/100 jLtL,  then further weighings are
      not necessary.  If the residue weight is greater  than 10 mg/100 /zL,


                            3640A - 12                        Revision 1
                                                          September 1994

-------
      then determine  if constant weight has been achieved by placing the
      weighing dish  and  residue back under the heat  lamp  for  2 or more
      additional  0.5  min.   intervals.    Reweigh  after each   interval.
      Constant weight  is achieved when three weights  agree within +10%.

            7.4.1.3     Repeat the above residue analysis on a  blank and
      a  spike.   Add  100 fj,l  of  the  same  methylene  chloride used for the
      sample extraction to a weighing dish and determine residue as above.
      Add 100 jitL of a corn oil spike (5 g/100 mL)  to another weighing dish
      and repeat the  residue determination.

      7.4.2 A residue weight of 10 mg/100 /zL of extract  represents 500 mg
in  5  ml  of extract.    Any  sample  extracts  that exceed  the  10  mg/100 /zL
residue weight must be diluted so that the  5 ml loaded  on the GPC column
does  not exceed  0.500 g.  When  making  the  dilution, keep in mind that a
minimum  volume  of  8  ml  is  required  when  loading  the  ABC  GPC  unit.
Following is a calculation that may  be  used  to determine what dilution is
necessary if the residue exceeds 10 mg.
      Y ml taken  =
      for dilution
Example:
      Y ml taken  =
      for dilution
10 ml final
  volume
10 ml final
  volume
 10 mq maximum
X mg of residue
 10 mq maximum
15 mg of residue
      Y ml taken for dilution  =  6.7 ml
      Therefore,  taking  6.7 ml of  sample  extract from  Sec.  7.3.2, and
diluting to 10 ml with methylene chloride,  will  result  in  5 ml of diluted
extract loaded on the GPC column that contains 0.500 g of residue.

      NOTE: This dilution factor must be included in the final calculation
            of analyte concentrations.  In the above example, the dilution
            factor is 1.5.

7.5   GPC Cleanup

      7.5.1 Calibrate  the  GPC  at  least   once  per  week following the
procedure outlined in Sees. 7.2.2 through 7.2.2.6.  Ensure that UV  trace
requirements,  flow  rate  and  column pressure  criteria  are  acceptable.
Also, the  retention  time shift must be  <5% when  compared  to retention
times in the last calibration UV trace.

            7.5.1.1     If these criteria  are not  met, try  cleaning the
      column by loading one or  more  5 ml portions  of butyl  chloride and
      running  it through  the  column.     Butyl  chloride or  9:1   (v/v)
      methylene   chloride/methanol   removes    the   discoloration   and
      particulate that may have precipitated out of the  methylene chloride
      extracts.   Backflushing  (reverse flow) with  methylene  chloride to
      dislodge  particulates  may restore lost  resolution.   If  a  guard
      column is being used, replace it  with a new one.   This may correct
                            3640A - 13
                                      Revision  1
                                  September  1994

-------
            the problem.   If column maintenance  does not  restore  acceptable
            performance, the  column  must be repacked  with  new Bio  Beads  and
            calibrated.

            7.5.2 Draw a minimum of 8 ml  of extract (diluted, if necessary,  and
      filtered) into a 10 ml syringe.

            7.5.3 Attach the  syringe to  the  turn  lock on the  injection  port.
      Use firm, continuous  pressure  to  push the  sample  onto the 5-mL  sample
      loop.   If the sample  is  difficult to load, some part of the system may be
      blocked.   Take  appropriate  corrective  action.   If the back pressure is
      normal  (6-10 psi),  the blockage is probably in the valve.  Blockage may be
      flushed out  of  the valve  by  reversing the inlet  and outlet  tubes  and
      pumping solvent through the tubes.   (This should be  done  before  sample
      loading.)

            NOTE:  Approximately  2 ml of the extract remains in the lines between
                  the injection port and the sample  loop;  excess sample also
                  passes through the  sample  loop to waste.

            7.5.4 After loading  a loop,  and before  removing the syringe from the
      injection port,  index the GPC to the next  loop.  This will  prevent loss of
      sample caused by unequal pressure in the  loops.

            7.5.5 After loading each sample  loop, wash the loading  port with
      methylene chloride in  a  PTFE wash bottle to minimize cross-contamination.
      Inject approximately  10 ml of methylene chloride  to rinse the  common
      tubes.

            7.5.6 After loading  all  the  sample loops,  index the  GPC  to  the 00
      position, switch  to  the  "RUN" mode and  start  the automated  sequence.
      Process each sample using  the collect and dump cycle times established in
      Sec.  7.2.2.

            7.5.7 Collect each sample in a 250 ml Erlenmeyer  flask, covered with
      aluminum foil to reduce solvent evaporation, or directly  into  a Kuderna-
      Danish evaporator.  Monitor sample  volumes collected.   Changes  in  sample
      volumes collected may indicate  one  or  more of the following problems:

                  7.5.7.1     Change  in solvent flow rate, caused by  channeling
            in the column or changes  in column  pressure.

                  7.5.7.2     Increase  in column operating pressure  due  to the
            absorption of particles or gel fines onto  either  the guard column or
            the analytical  column gel,  if a  guard  column  is  not used.

                  7.5.7.3     Leaks  in  the system or  significant  variances in
            room temperature.

      7.6   Concentrate the extract by the standard K-D technique (see  any  of the
extraction  methods, Sec.  4.2.1 of this chapter).  See the determinative methods
(Chapter Four,  Sec. 4.3) for the final  volume.
                                  3640A - 14                        Revision 1
                                                                September 1994

-------
      7.7   It  should  be remembered that  only  half of the  sample extract is
processed by the GPC (5 ml of the 10 ml  extract  is loaded onto the  GPC column),
and thus, a dilution factor of 2  (or 2 multiplied by  any dilution  factor  in Sec.
7.4.2) must be used for quantitation of  the sample in the determinative method.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One and Method 3600 for specific quality control
procedures.

      8.2   The analyst should demonstrate that  the  compound(s) of  interest are
being quantitatively recovered before applying this method to actual samples.

      8.3   For  sample extracts  that  are cleaned  up  using this  method,  the
associated quality control samples must also be processed through  this cleanup
method.
9.0   METHOD PERFORMANCE

      9.1   Refer to Table 1 for single laboratory performance data.


10.0  REFERENCES

1.    Wise, R.H.; Bishop, D.F.; Williams, R.T.; Austern, B.M.  "Gel Permeation
      Chromatography in  the  GC/MS  Analysis of Organics  in  Sludges";  U.S.  EPA
      Municipal Environmental Research Laboratory:  Cincinnati, Ohio 45268.

2.    Czuczwa,  J.; Alford-Stevens, A.  "Optimized  Gel Permeation Chromatographic
      Cleanup for Soil,  Sediment, Waste and Waste  Oil Sample  Extracts for GC/MS
      Determination of Semivolatile Organic Pollutants, JAOAC, submitted April
      1989.

3.    Marsden,  P.J.; Taylor, V.; Kennedy, M.R.  "Evaluation of Method 3640 Gel
      Permeation  Cleanup";   Contract  No.   68-03-3375,   U.S.   Environmental
      Protection Agency, Cincinnati, Ohio, pp. 100,  1987.
                                  3640A -  15                        Revision 1
                                                                September 1994

-------
                  TABLE 1
GPC RECOVERY AND RETENTION VOLUMES FOR RCRA
          APPENDIX VIII  ANALYTES
Compound
Acenaphthene
Acenaphthylene
Acetophenone
2-Acetylaminofluorene
Aldn'n
4-Aminobiphenyl
Aniline
Anthracene
Benomyl
Benzenethiol
Benzidine
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Benzo(k)fluoranthene
Benzoic acid
Benzotrichloride
Benzyl alcohol
Benzyl chloride
alpha-BHC
beta-BHC
gamma- BHC
delta-BHC
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
2-sec-butyl-4,6-dinitrophenol (Dinoseb)
Carbazole
Carbendazim
alpha-Chlordane
gamma-Chlordane
4-Chloro-3-methyl phenol
4-Chloroaniline
Chi orobenzi late
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
2-Chloronaphthalene
2-Chlorophenol
3-Chlorophenol
4-Chlorophenol
4-Chlorophenyl phenyl ether
3-Chloropropionitrile
Chrysene
2-Cresol
% Rec1
97
72
94
97
99
96
93
89
131
92
95
100
93
93
90
91
66
93
95
99
84
94
93
102
93
104
103
99
131
97
93
87
88
92
89
76
83
89
90
86
87
98
80
102
91
% RSD2
2
10
7
2
9
7
4
2
8
11
5
3
5
3
6
4
7
7
17
4
13
9
4
7
1
3
18
5
8
2
2
1
3
5
1
2
2
1
1
3
2
2
5
1
1
Ret. Vol.3 (ml
196-235
196-235
176-215
156-195
196-215
176-215
196-235
196-235
146-195
196-235
176-215
196-235
196-235
196-235
196-235
196-235
176-195
176-215
176-215
176-215
196-215
196-215
196-215
216-255
176-215
136-175
176-195
196-255
146-195
196-235
196-215
196-255
196-235
176-235
156-195
156-215
156-195
196-235
196-215
196-215
196-215
176-215
176-215
196-235
196-215
                 3640A -  16
    Revision 1
September 1994

-------
TABLE 1 (continued)
Compound
3-Cresol
4-Cresol
Cyclophosphamide
ODD
DDE
DDT
Di-n-butyl phthalate
Dial! ate
Dibenzo(a,e)pyrene
Dibenzo(a,i)pyrene
Dibenz(a, j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Dibenzothiophene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
trans-l,4-Dichloro-2-butene
cis-l,4-Dichloro-2-butene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 ,4-Dichlorobenzene
3,3'-Dichlorobenzidine
2,6-Dichlorophenol
2,4-Dichlorophenoxyacetic acid (2,4-D)
2,4-Dichlorophenol
2,4-Dichlorotoluene
l,3-Dichloro-2-propanol
Dieldrin
Diethyl phthalate
Dimethoate
3,3'-Dimethoxybenzidinea
Dimethyl phthalate
p-Dimethylaminoazobenzene
7, 12-Dimethyl -benz(a)anthracene
2,4-Dimethylphenol
3, 3' -Dimethyl benzi dine
4,6-Dinitro-o-cresol
1,3-Dinitrobenzene
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diphenylamine
Diphenyl ether
1 , 2-Di phenyl hydrazine
Disulfoton
Endosulfan sulfate
Endosulfan I
% Rec1
70
88
114
94
94
96
104
97
94
99
117
92
94
94
83
121
107
106
81
81
81
98
86
80
87
70
73
100
103
79
15
100
96
77
93
93
100
99
118
93
101
95
67
92
81
94
99
%RSD2
3
2
10
4
2
6
3
6
10
8
9
5
1
3
2
8
6
6
1
1
1
3
3
NA
2
9
13
5
3
15
11
1
1
1
2
2
1
2
7
4
2
6
12
1
15
2
8
Ret. Vol.3 (ml
196-215
196-215
146-185
196-235
196-235
176-215
136-175
156-175
216-235
216-235
176-195
196-235
176-235
196-235
176-215
196-215
176-195
176-215
196-235
196-235
196-235
176-215
196-215
76-215
96-215
196-235
176-215
196-215
136-195
146-185
156-195
156-195
176-215
176-215
176-215
156-215
156-195
156-195
176-195
156-195
156-175
176-235
196-215
176-215
146-165
176-195
176-215
    3640A - 17
    Revision 1
September 1994

-------
TABLE 1 (continued)
Compound
Endosulfan II
Endrin
Endrin aldehyde
Endrin ketone
Ethyl methane sulfonate
Ethyl methacrylate
Bis(2-ethylhexyl) phthalate
Famphur
Fluorene
Fluoranthene
Heptachlor
Heptachlor epoxide
Hexachl orobenzene
Hexachlorobutadiene
Hexachl orocycl opentadi ene
Hexachl oroethane
Hexachl oropropene
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
cis-Isosafrole
trans-Isosafrole
Kepone
Malononitrile
Merphos
Methoxychlor
3-Methylchol anthrene
2-Methyl naphthal ene
Methyl parathion
4,4'-Methylene-bis(2-chloroaniline)
Naphthalene
1,4-Naphthoquinone
2-Naphthylamine
1-Naphthylamine
5-Nitro-o-toluidine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
2-Nitrophenol
4-Nitrophenol
N-Nitroso-di-n-butylamine
N-Nitrosodiethanolamine
N-Nitrosodi ethyl ami ne
N-Nitrosodimethylamine
N-Nitrosodiphenylamine
N-Nitrosodi-n-propylamine
% Rec1
92
95
97
94
62
126
101
99
95
94
85
91
108
86
89
85
91
79
98
68
90
88
102
111
93
94
74
67
84
96
95
73
94
96
77
96
96
103
86
95
77
89
104
94
86
99
85
%RSD2
6
6
1
4
7
7
1
NA
1
1
2
11
2
2
3
1
2
13
5
7
4
16
NA
9
12
6
12
6
13
1
7
7
8
6
2
8
2
8
2
3
3
4
3
2
13
2
4
Ret. Vol.3 (ml
196-215
196-215
176-215
176-215
176-235
176-195
120-145
126-165
176-235
196-235
195-215
156-195
196-235
176-215
176-215
196-235
196-235
216-255
196-235
156-195
176-215
156-195
196-235
156-195
126-165
156-195
176-195
196-215
146-185
176-215
196-215
176-215
196-235
196-235
176-195
176-215
176-215
176-215
176-195
176-195
196-215
156-175
146-185
156-175
156-195
156-195
156-175
     3640A - 18
    Revision 1
September 1994

-------
                                 TABLE 1 (continued)
Compound
N-Ni trosomethyl ethyl ami ne
N-Nitrosomorphol ine
N-Nitrosopiperidine
N-Nitrosopyrol idine
Di-n-octyl phthalate
Parathion
Pentachl orobenzene
Pentachl oroethane
Pentachl oronitrobenzene (PCNB)
Pentachl orophenol
Phenacetin
Phenanthrene
Phenol
1,2-Phenylenediamine
Phorate
2-Picoline
Pronamide
Pyrene
Resorcinol
Safrole
Streptozotocin8
1,2, 4, 5-Tetrachl orobenzene
2, 3, 5, 6-Tetrachloro -nitrobenzene
2,3,4 , 6-Tetrachl orophenol
2,3,5,6-Tetrachlorophenol
Tetraethyl dithiopyrophosphate (Sulfotep)
Thiosemicarbazide
2-Toluidine
4-Toluidine
Thiourea, l-(o-chlorophenyl)
Toluene-2,4-diamine
1,2, 3 -Trichl orobenzene
1,2, 4 -Trichl orobenzene
2, 4, 5-Trichl orophenol
2, 4, 6-Trichl orophenol
2,4,5-Trichlorophenoxyacetic acid (2,4,5-T)
2,4,5-Trichlorophenoxypropionic acid
Warfarin
% Rec1
83
86
84
92
83
109
95
74
91
102
100
94
83
91
74
99
105
98
70
93
6
96
85
95
96
89
74
92
87
75
69
87
89
77
95
71
67
94
%RSD2
7
4
4
1
4
14
2
1
8
1
3
2
2
1
NA
14
15
2
6
1
48
2
9
1
7
14
3
3
8
11
7
1
1
1
1
23
NA
2
Ret. Vol.3 (ml
156-175
156-195
156-195
156-175
120-156
146-170
196-235
196-235
156-195
196-215
156-195
196-235
156-195
196-215
116-135
156-215
156-195
215-235
196-215
176-215
225-245
196-235
176-215
196-215
196-215
116-135
146-185
176-235
176-235
166-185
176-215
196-235
196-235
216-235
216-235
156-235
216-215
166-185
NA = Not applicable, recovery presented as the average of two determinations.

a  Not an appropriate analyte for this method.

1  The percent recovery is based on an average of three recovery values.

2  The % relative standard deviation is determined from three recovery values.

3  These Retention Volumes are  for guidance  only  as  they  will  differ from column to
   column and from system to system.
                                     3640A - 19
    Revision 1
September 1994

-------
                         Figure 1
        GPC RETENTION VOLUME OF CLASSES OF ANALYTES
                                  W//////////////,
                                  W/////////////A
        PHTHALATB -»•
onOANOPHOSPHATE
    PESTICIDES
   CORN
                                  PAH'*
                                  CHLOROBENZENES
                       NITR03AMINE3, NITROAROMATICS
                                  AROMATIC AMINES
                           NITROPHCNOL3
                               CHLOROPHENOL3
                                  OROANOCHLORINE
                                  PESTICIDES/PCS'*
                               HERBICIDES 
-------
                              Figure 2
            UV CHROMATOGRAM OF THE CALIBRATION SOLUTION
Injection
5 fflls
on column
                                                               — 0 minaces
Corn oil
25 mg/nL
Bis(2-ethylhesyl) -phthaiate
1.0 mg/niL
Methoxychlor
0.2 ng/mL
Perylene
0.02 mg/mL
Sulfur
0.08 mg/oL  —  ","~
                                                             15 minuces
                                                           "' 30 minutes
                                                                   45 minutes
700 am X25 am col-
70 g Bio-Beads SX
Bed length » 490
CH.C12  at 5.0 uL
254 nu
                                  1-1.:	~	"1_ "_:.._"......"_! 60 minutes
            *  f _, .
            mn« __
                              3640A - 21
                                                                      Revision 1
                                                                  September 1994

-------
                                              METHOD  3640A
                                       GEL-PERMEATION  CLEANUP
                               7 1 Ensure ambient temp, consistent
                                      throughout GPC run.
                                 7.2 GPC Setup and Calibration
      7 2.1 Column Preparation
               I
  7.2.1.1 Place Bio Beads and MeCI
      in a container. Swirl and
        allow beads to swell.
                I
    7 2.1.2 Remove column inlet bed
  support plunger.  Position and tighten
outlet bed support plunger to column end.
   7 2.1 3 Ensure GPC column outlet
 contains solvent. Place small amount
     solvent in column to minimize
          bubble formation.
   7.2.1.4 Transfer bead mixture into
   sep. funnel. Drain excess solvent;
    drain beads into column. Keep
        beads wet throughout.
    7.2 1 5 Loosen seal on opposite
  plunger assembly, insert into column.
                                                                                       1
 7 2.1.6 Compress column  Slurry
remaining beads and repeat Section
 7 2.1.5 and column compression.
   7.2.1.7 Compress column bed
      approximately four cm.
  7 2.1.8 Pack option 5 cm. guard
     column w/ roughly 5 gm.
 	preswelled beads	
   7 2.1.9 Connect column inlet to
   solvent reservoir. Pump MeCI at
         5 ml/mm for 1 hr
                                                                                       I
  7.2.1.10 Connect column outlet to
  UV-Vis detector  Place restrictor
  at detector outlet. Run MeCI for
   additional 1 -2 hrs.  Compress
  column bed to provide 6-10 psi
          backpressure.
                                                                         7.2.1.11 Connect outlet line to column
                                                                         inlet when column not in use. Repack
                                                                         column when channeling is observed.
                                                                         Assure consistent backpressure when
                                                                           beads are rewetted after drying.
                                                3640A  -  22
                        Revision  1
                   September 1994

-------
                                        METHOD  3640A
                                          continued
  7 2.2 Calibration of the GPC column
     7 2.2.1 Load sample loop with
          calibration solution.
                I
  7 2.2.2 Inject calibration soln.; adjust
     recorder or detector sensitivity
  to produce similar UV trace as Fig 2
     7 2.2.3 Evaluation criteria for
         UV chromatogram.
  7.2 2.4 Calibration for Semivolatiles
    Use information from UV trace to
     obtain collect and dump times.
Initiate collection before bis(2-ethylhexyl)
 phthalate, stop after perylene. Stop run
          before sulfur e/utes.
 7 2.2.5 Calibration for Organochlonne
          Pesticides/PCBs
  Choose dump time which removes
    > 85% phthalate, but collects at
   times > 95% methoxychlor  Stop
   collection between perylene and
           sulfur elution
               I
   7 2.2.6 Verify column flow rate and
        backpressure  Correct
     inconsistencies when criteria
            are not met.
7.2.2.7 Reinject calibration soln when
   collect and dump cycles are set,
    and column criteria are met
                                                                      7 2.2.7 1 Measure and record
                                                                         volume of GPC eluate
 7.2.2.7 2 Correct for retention time
       shifts of > W- 5% for
    bis(2-ethylhexyl) phthalate
          and perylene.
 7.2.2.8 Inject and analyze GPC blank
    for column cleanliness  Pump
   through MeCI as column wash.
                                         3640A  -  23
                           Revision   1
                    September  1994

-------
METHOD 3640A
  continued
V
7 3 Extract Preparation
*
7.3.1 Adjust extract volume to 10 ml.
Primary solvent should be MeCI.
*
7.3.2 Filter extract through 5 micron filter
disc/syringe assembly into small
glass container.
*
7.4 Screening the Extract
*
7.4.1 Screen extract by determining
residue wt. of 100 uL aliquot.
*
7.4.1 .1 Transfer 100 uL of filtered
extract from Section 7.3.2 to tared
aluminum weighing dish.
*
7.4.1 .2 Evaporate extract solvent under
heating lamp. Weigh residue to nearest
0 1 mg.
*
7.4.1 .3 Repeat residue analysis of Section
7412 w/blank and spike sample.
*
7 .4 2 Use dilution example to determine
necessary dilution when residue
wts. > 10mg.




1 '
7.5 GPC Cleanup
*
7.5.1 Calibrate GPC weekly. Assure
column criteria. UV trace, retention
time shift criteria are met.
*
7.5.1 .1 Clean column w/butyl chloride
loadings, or replacement of
guard column.
*
7.5.2 Draw 8 ml extract into syringe.
*
7.5.3 Load sample into injection loop.
I
7.5.4 Index GPC to next loop to
prevent sample loss.
1
7.5.5 Wash sample port w/MeCI
between sample loadings.
*
7 5.6 At end of loadings, index GPC to
00, switch to "RUN" mode, start
automated sequence.
*
7.5 7 Collect sample into aluminum foil
covered Erlenmeyer flask or into
Kuderna- Danish evaporator
i
7.6 Concentrate extract by std.
Kuderna-Danish technique.
*
7.7 Note dilution factor of GPC method
into final determinations.
 3640A - 24
    Revision 1
September 1994

-------
                                 METHOD 3650A

                          ACID-BASE  PARTITION  CLEANUP
1.0   SCOPE AND APPLICATION
manual.
      1.1   Method 3650 was formerly Method 3530 in the second edition of this
      1.2   Method  3650  is  a  liquid-liquid  partitioning  cleanup method  to
separate  acid analytes,  e.g.   organic  acids  and  phenols,  from  base/neutral
analytes, e.g. amines, aromatic  hydrocarbons, and halogenated  organic compounds,
using pH  adjustment.   It  may be used for cleanup of  petroleum waste  prior to
analysis or further cleanup (e.g., alumina cleanup).  The following compounds can
be separated by this method:
Compound Name
CAS No.a
Fraction
Benz ( a ) anthracene
Benzo(a)pyrene
Benzo(b)fluoranthene
Chlordane
Chlorinated dibenzodioxins
2-Chlorophenol
Chrysene
Creosote
Cresol(s)
Dichlorobenzene(s)
Dichlorophenoxyacetic acid
2, 4-Dimethyl phenol
Dinitrobenzene
4,6-Dinitro-o-cresol
2,4-Dinitrotoluene
Heptachlor
Hexachlorobenzene
Hexachl orobutadi ene
Hexachloroethane
Hexachl orocyclopentadi ene
Naphthalene
Nitrobenzene
4-Nitrophenol
Pentachlorophenol
Phenol
Phorate
2-Picoline
Pyridine
Tetrachlorobenzene(s)
Tetrachlorophenol (s)
Toxaphene
Trichlorophenol(s)
2,4,5-TP (Silvex)
56-55-3
50-32-8
205-99-2
57-74-9

95-57-8
218-01-9
8001-58-9


94-75-7
105-67-9
25154-54-5
534-52-1
121-14-2
76-44-8
118-74-1
87-68-3
67 -.72-1
77-47-4
91-20-3
98-95-3
100-02-7
87-86-5
108-95-2
298-02-2
109-06-8
110-86-1


8001-35-2

93-72-1
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Acid
Base-neutral
Base-neutral and Acid
Acid
Base-neutral
Acid
Acid
Base-neutral
Acid
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Acid
Acid
Acid
Base-neutral
Base-neutral
Base-neutral
Base-neutral
Acid
Base-neutral
Acid
Acid
   Chemical  Abstract Services Registry Number.

                                   3650A  -  1
                                   Revision 1
                                    July 1992

-------
2.0   SUMMARY OF METHOD

      2.1   The solvent extract from a prior solvent extraction method is shaken
with water that  is  strongly basic.  The acid analytes partition into the aqueous
layer, whereas,  the basic and neutral compounds stay in the organic solvent.  The
base/neutral fraction is concentrated  and  is then ready for further cleanup, if
necessary, or analysis.  The aqueous  layer is acidified  and  extracted  with an
organic solvent.  This extract  is concentrated (if necessary) and is then ready
for analysis of the acid analytes.


3.0   INTERFERENCES

      3.1   More extensive procedures  than those outlined in this method may be
necessary for reagent purification.

      3.2   A method  blank must  be run  for the  compounds of  interest prior to
use of the method.  The interferences  must be below the method detection limit
before this method is applied to actual  samples.


4.0   APPARATUS AND MATERIALS

      4.1   Drying column  -  20 mm ID Pyrex chromatographic  column  with Pyrex
glass wool at bottom, or equivalent.

      NOTE: Fritted glass discs are difficult to clean  after highly contaminated
            extracts  have been passed through them.  Columns without frits are
            recommended.   Use  a  small  pad of  Pyrex glass wool to  retain the
            adsorbent.   Prewash  the  glass  wool  pad  with  50  ml   of  acetone
            followed  by 50  ml  of elution  solvent prior  to packing  the column
            with adsorbent.

      4.2   Kuderna-Danish (K-D)  apparatus

            4.2.1  Concentrator tube -  10  ml graduated (Kontes K570050-1025 or
      equivalent).  A ground glass stopper  is used  to prevent evaporation of the
      extracts.

            4.2.2  Evaporation  flask  -  500 ml  (K-570001-0500  or  equivalent).
      Attach to concentrator tube with  springs, clamps,  or equivalent.

            4.2.3  Snyder  column  - Three  ball  macro  (Kontes  K-503000-0121 or
      equivalent).
            4.2.4  Snyder  column  - Two ball  micro  (Kontes  K569001-0219  or
      equivalent).
tops.
            4.2.5  Springs  -  1/2  inch  (Kontes  K-662750  or equivalent).

      4.3   Vials  - Glass, 2 ml capacity with Teflon lined screw-caps or crimp
      4.4   Water  bath  - Heated, concentric ring cover, temperature control of
± 2°C.  Use this bath in a hood.

                                   3650A -  2                         Revision 1
                                                                     July 1992

-------
      4.5   Boiling chips -  Solvent extracted, approximately 10/40 mesh (silicon
carbide or equivalent).

      4.6   pH indicator paper - pH range including the desired extraction pH.

      4.7   Separatory funnel - 125 mi.

      4.8   Erlenmeyer flask  - 125 ml.


5.0   REAGENTS

      5.1   Reagent  grade  inorganic  chemicals  shall  be  used  in  all  tests.
Unless otherwise  indicated,  it  is intended that  all  inorganic  reagents shall
conform to the specifications of  the Committee on Analytical Reagents  of the
American Chemical  Society, where such specifications are available.  Other grades
may be used, provided it is first ascertained that the reagent is of sufficiently
high  purity  to  permit  its  use  without  lessening  the  accuracy  of  the
determination.

      5.2   Organic-free reagent water - All  references  to  water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Sodium hydroxide, NaOH,  (ION)  -  Dissolve 40 g of sodium hydroxide
in 100 ml of organic-free reagent water.

      5.4   Sulfuric acid, H?S04,  (1:1 v/v in water) - Slowly add 50 ml H2S04 to
50 ml of organic-free reagent water.

      5.5   Sodium sulfate  (granular, anhydrous), Na2SO,  -  Purify by heating at
400°C for 4 hours in  a shallow tray, or by precleaning the sodium sulfate with
methylene chloride.  If the sodium  sulfate is precleaned with methylene chloride,
a method blank must be analyzed, demonstrating that there is no interference from
the sodium sulfate.

      5.6   Solvents:

            5.6.1 Methylene chloride, CH2C12  -  Pesticide  quality or equivalent.

            5.6.2 Acetone,  CH3COCH3 - Pesticide quality  or equivalent.

            5.6.3 Methanol, CH3OH - Pesticide quality or equivalent.

            5.6.4 Diethyl  Ether,  C2H5OC2H5 - Pesticide  quality  or equivalent.
      Must  be free of  peroxides  as  indicated by test  strips  (EM  Quant,  or
      equivalent).  Procedures for  removal of  peroxides are provided with the
      test strips.  After cleanup, 20 ml of ethyl alcohol preservative must be
      added to each liter of  ether.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory  material to  this chapter,  Organic Analytes,
Section 4.1.

                                   3650A - 3                         Revision 1
                                                                     July 1992

-------
7.0   PROCEDURE

      7.1   Place 10 ml of the solvent extract from a prior extraction procedure
into a 125 ml separatory funnel.

      7.2   Add 20 mL of methylene chloride to the separatory funnel.

      7.3   Slowly add 20 ml of prechilled  organic-free reagent water which has
been previously adjusted to a pH of 12-13 with ION sodium hydroxide.

      7.4   Seal and  shake the separatory funnel for at least  2  minutes with
periodic venting to release excess pressure.

      NOTE: Methylene   chloride   creates   excessive   pressure   very  rapidly;
            therefore,  initial  venting should be done immediately  after the
            separatory funnel has been sealed and shaken once.   The separatory
            funnel should be vented into a  hood to prevent unnecessary exposure
            of the analyst to the organic  vapor.

      7.5   Allow the  organic  layer  to separate  from the aqueous  phase for a
minimum of 10 minutes.   If the  emulsion  interface between layers  is more than
one-third the  size  of the solvent layer,  the analyst must  employ mechanical
techniques to complete the  phase separation.  The  optimum technique depends upon
the sample, and may include stirring, filtration  of the emulsion through glass
wool, centrifugation, or other physical methods.

      7.6   Separate the  aqueous  phase  and transfer  it  to a  125 ml Erlenmeyer
flask.   Repeat  the  extraction two more  times using  20 ml aliquots of dilute
sodium hydroxide (pH 12-13).  Combine the aqueous extracts.

      7.7   Water soluble  organic acids and phenols will be primarily in the
aqueous phase.  Base/neutral  analytes will be in the methylene chloride.   If the
analytes  of  interest are  only in the aqueous  phase,  discard the methylene
chloride and proceed  to Section 7.8.  If the analytes of interest are only in the
methylene chloride,  discard the aqueous phase and proceed to Section 7.10.

      7.8   Externally cool the 125 mL Erlenmeyer flask with ice while adjusting
the  aqueous  phase  to a  pH of 1-2 with  sulfuric acid  (1:1).   Quantitatively
transfer the cool  aqueous phase to a  clean  125 mL  separatory funnel.  Add 20 mL
of methylene chloride to the  separatory funnel and shake for at least 2 minutes.
Allow the methylene  chloride to separate from  the  aqueous phase and collect the
methylene chloride in an Erlenmeyer flask.

      7.9   Add 20 mL of methylene chloride to the separatory funnel  and extract
at pH 1-2 a second time.  Perform a third extraction in the same manner combining
the extracts in the Erlenmeyer flask.

      7.10  Assemble  a  Kuderna-Danish  (K-D) concentrator  (if necessary)  by
attaching a 10 mL concentrator tube to a 500 mL evaporation  flask.

      7.11  Dry both acid  and base/neutral fractions by passing them through a
drying column containing about 10 cm of anhydrous sodium sulfate.   Collect the
dried  fractions in  K-D  concentrators.    Rinse   the  Erlenmeyer  flasks which

                                   3650A -  4                        Revision 1
                                                                     July 1992

-------
contained the  solvents  and the  columns  with 20  ml  of methylene  chloride to
complete the quantitative transfer.

      7.12  Concentrate both  acid  and  base/neutral fractions  as follows:  Add
one or  two  boiling chips  to  the flask  and  attach  a three ball  macro-Snyder
column.  Prewet the Snyder column by  adding about  1 ml of methylene chloride to
the top of the  column.  Place  the K-D apparatus on a hot  water  bath  (80-90°C) so
that the concentrator tube is  partially immersed  in the warm water.  Adjust the
vertical position  of  the apparatus  and  the  water temperature  as  required to
complete the concentration in  15-20 minutes.  At the proper rate of distillation,
the balls of the column will  actively chatter but the chambers will not flood.
When the apparent volume of liquid reaches 1  ml,  remove the K-D apparatus from
the water bath  and  allow  it to cool.  Remove the Snyder column and rinse the
flask and its lower joints into the concentrator tube with 1-2 ml of methylene
chloride.  Concentrate the extract to the final  volume using either the micro-
Snyder column technique (7.12.1) or nitrogen  blowdown technique (7.12.2).

            7.12.1      Micro-Snyder Column  Technique

                   7.12.1.1    Add  another one  or two  boiling  chips  to  the
            concentrator tube and attach  a two ball micro-Snyder column.  Prewet
            the column by adding  0.5 ml  of methylene chloride  to the top of the
            column.   Place the K-D apparatus in  a hot water  bath  (80-90°C) so
            that the concentrator tube is partially immersed  in the hot water.
            Adjust  the  vertical  position  of  the   apparatus  and the water
            temperature  as required  to  complete the  concentration   in  5-10
            minutes.  At the proper rate  of distillation the balls of the column
            will actively chatter but  the chambers  will not  flood.   When the
            apparent  volume  of  the liquid  reaches  0.5  ml,  remove   the  K-D
            apparatus and allow it to cool.   Remove the  Snyder column and rinse
            the flask and  its lower joints into the concentrator tube with 0.2
            ml  of  methylene  chloride.    Adjust the  final  volume to 1  ml with
            methylene chloride.

            7.12.2      Nitrogen Blowdown Technique

                   7.12.2.1     Place the  concentrator tube in  a warm water bath
            (35°C)  and evaporate the  solvent volume to 1.0-2.0  ml using  a gentle
            stream  of  clean,  dry  nitrogen  (filtered  through  a column  of
            activated carbon).

                   CAUTION:     Do not use plasticized tubing between the carbon
                               trap and the sample.

                   7.12.2.2    The internal wall of the concentrator tube must be
            rinsed  down several  times  with the  appropriate solvent during the
            operation.   During  evaporation,  the tube  solvent level   must be
            positioned to avoid  condensation  water.   Under normal procedures,
            the extract must  not  be  allowed to become dry.

                   CAUTION:     When the  volume of  solvent is  reduced below 1 ml,
                               semivolatile analytes  may be lost.

      7.13  The acid  fraction is now ready for  analysis.   If the base/neutral

                                   3650A  - 5                         Revision 1
                                                                     July 1992

-------
fraction requires further cleanup by  the  alumina  column cleanup for petroleum
waste (Method 3611),  the solvent may have to be changed to hexane.  If a solvent
exchange is required, momentarily  remove the Snyder column, add approximately 5
ml of  the  exchange  solvent  and  a new boiling  chip,  and reattach  the Snyder
column.  Concentrate the extract as described in Section 7.12.1.1, raising the
temperature of the water bath,  if necessary,  to maintain proper distillation.
When the apparent volume again reaches 1  ml,  remove the K-D apparatus from the
water bath and allow it to drain and cool  for at least 10 minutes.  Repeat the
exchange 2 more  times.   If no further cleanup  of the base/neutral  extract is
required, it is also ready for analysis.


8.0   QUALITY CONTROL

      8.1   Refer  to Chapter One for general  quality  control  procedures and
Method 3600 for cleanup procedures.

      8.2   The  analyst must demonstrate  that  the  compounds  of  interest are
being quantitatively recovered before applying this method to actual samples.

      8.3   For  samples that  are cleaned using  this method,  the  associated
quality control samples must be processed through this cleanup method.


9.0   METHOD PERFORMANCE

      9.1   Refer to the determinative methods  for performance data.


10.0  REFERENCES

1.    Test Methods;   Methods  for Organic Chemical  Analysis  of Municipal and
       Industrial Wastewater;   U.S.  Environmental  Protection Agency. Office of
      Research and Development. Environmental Monitoring and  Support Laboratory.
      ORD Publication Offices of Center for Environmental  Research Information:
      Cincinnati, OH, 1982;  EPA-600/4-82-057.
                                   3650A - 6                         Revision  1
                                                                      July  1992

-------
                                                  METHOD 3650A
                                       ACID-BASE  PARTITION  CLEANUP
      START
 7  1  Place extract
 or organic liquid
    waste into
 sapara to ry funnel
 7  2 Add mathylena
    chlorxde
7  3  Add prechilled
  dilute 3odium
     hydroxide
7 4  Seal and shake
 separatory funnel
     7 5 Allow
   separation of
or gam c layer from
   aqueous phase

— >
7 5 Complete phase
mechanical
techniques
                               7  6  Transfer
                             aqueous phase to
                               fJasV, repeat
                             extraction twice,
                              combine aqueous
                                 ex t racts
7  7  Discard aqueous
       phase
                                                                      Aqueous
                             7 10  Assemolc
                                apparaLus
                                          K-D
7  7  Discard organic
      phase
                                                                            7  8  Adjust pH with
                                                                           JulJfuric acid,  trans
                                                                           fer aqueous phase to
                                                                           clean scparatory fun
                                                                            nel . add ntethylene
                                                                             chloride.  shake,
                                                                            allow phase separa-
                                                                           tion, collee t solven
                                                                              phase in flask
7 9  Perform 2 mote
  ex tractions,
    combine all
    extracts
                                                     3650A  -  7
                                                                            Revision  1
                                                                             July  1992

-------
         METHOD 3650A
          (Continued)
711 Dry extracts,
collect extracts in
 K-D concen t ra t o r,
 rinse flask with
methylene chloride
 7  12 Concent rate
  both f tac Lions
                          ' 14  LKchange
                            solvent
     Ana 1y ze
   fractions by
    appropriate
   determinative
     me t h o d
           3650A - 8
Revision  1
 July 1992

-------
                                 METHOD 3660A

                                SULFUR CLEANUP
1.0   SCOPE AND APPLICATION

      1.1   Elemental sulfur is encountered  in many sediment samples (generally
specific to different areas in the country), marine algae, and some industrial
wastes.  The  solubility  of sulfur in various solvents is  very  similar to the
organochlorine  and  organophosphorus   pesticides.     Therefore,   the  sulfur
interference follows along with the pesticides through the  normal extraction and
cleanup techniques.  In general,  sulfur  will usually elute entirely in Fraction
1 of the Florisil cleanup  (Method 3620).

      1.2   Sulfur  will  be quite evident  in  gas chromatograms  obtained from
electron capture detectors, flame photometric detectors operated in the sulfur
or phosphorous  mode, and  Coulson  electrolytic  conductivity detectors  in the
sulfur mode.  If the gas  chromatograph is operated at the normal conditions for
pesticide analysis, the sulfur interference  can completely mask the region from
the solvent peak through  Aldrin.

      1.3   Three techniques for the elimination of sulfur are detailed within
this method: (1) the use  of copper powder;  (2) the use of mercury; and (3) the
use of tetrabutylammonium sulfite.  Tetrabutylammonium sulfite causes the least
amount of degradation of a broad range of pesticides and organic compounds, while
copper  and  mercury may  degrade  organophosphorus  and  some  organochlorine
pesticides.


2.0   SUMMARY OF METHOD

      2.1   The sample  to undergo cleanup is mixed with either copper, mercury,
or tetrabutylammonium (TBA) sulfite.   The mixture is shaken and the extract is
removed from the sulfur cleanup reagent.


3.0   INTERFERENCES

      3.1   Removal of sulfur using copper:

            3.1.1 The copper must be very reactive.   Therefore,  all  oxides of
      copper must be removed so that the copper has a shiny, bright appearance.

            3.1.2 The  sample   extract must  be  vigorously  agitated  with  the
      reactive copper for at least one minute.


4.0   APPARATUS AND MATERIALS

      4.1   Mechanical  shaker or mixer - Vortex  Genie or  equivalent.

      4.2   Pipets, disposable - Pasteur type.


                                  3660A - 1                         Revision 1
                                                                     July 1992

-------
      4.3   Centrifuge tubes, calibrated - 12 ml.

      4.4   Glass bottles or vials -  10 mL  and  50  mL,  with Teflon-lined screw
caps or crimp tops.

      4.5   Kuderna-Danish  (K-D) apparatus.

            4.5.1 Concentrator tube -   10 ml graduated  (Kontes K-570050-1025 or
      equivalent).  A  ground  glass stopper is used to prevent  evaporation of
      extracts.

            4.5.2 Evaporation   flask  -    500 ml  (Kontes   K-570001-500   or
      equivalent).   Attach  to  concentrator tube  with  springs,  clamps,  or
      equivalent.

            4.5.3 Snyder column  -   Three ball macro  (Kontes  K-503000-0121 or
      equivalent).

            4.5.4 Snyder column  -   Two ball micro  (Kontes  K-569001-0219 or
      equivalent).

            4.5.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform  to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are available. Other grades may be used, provided it is first
ascertained that  the reagent  is  of sufficiently high  purity to permit its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Nitric acid, HN03, dilute.

      5.4   Solvents

            5.4.1 Acetone, CH3COCH3  -  Pesticide quality or equivalent.

            5.4.2 Hexane, C6HU -  Pesticide quality or  equivalent.

            5.4.3 2-Propanol, CH3CH(OH)CH3 -  Pesticide  quality or  equivalent.

      5.5   Copper powder - Remove oxides by  treating  with dilute nitric acid,
rinse with organic-free reagent water to remove all traces of acid, rinse with
acetone and dry under a stream of nitrogen.  (Copper, fine  granular Mallinckrodt
4649 or equivalent).

      5.6   Mercury, triple distilled.

      5.7   Tetrabutylammonium (TBA) sulfite  reagent

                                   3660A - 2                        Revision  1
                                                                      July  1992

-------
            5.7.1 Tetrabutylammonium hydrogen sulfate,  [CH3(CH2)3]4NHS04.

            5.7.2 Sodium sulfite, Na2S03.

            5.7.3 Prepare  reagent  by  dissolving  3.39  g  tetrabutylammonium
      hydrogen  sulfate  in  100   ml organic-free  reagent  water.  To  remove
      impurities, extract  this solution  three  times with  20 ml  portions  of
      hexane.  Discard the hexane extracts,  and add 25 g sodium sulfite to the
      water solution.   Store  the resulting  solution, which  is  saturated with
      sodium sulfite, in an amber bottle  with a Teflon-lined screw cap.   This
      solution can be stored at room temperature for at least one month.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory  material to this chapter,  Organic Analytes,
Section 4.1.


7.0   PROCEDURE

      7.1   Removal  of sulfur using copper

            7.1.1 Concentrate  the  sample to exactly  1.0 mL  or  other  known
      volume.  Perform  concentration using  the  Kuderna-Danish (K-D) Technique
      (Method 3510,  Sections 7.10.1 through  7.10.4).

            CAUTION;    When  the  volume  of solvent  is  reduced  below  1  mL,
                        semi volatile analytes may be lost.

            7.1.2 If  the  sulfur  concentration  is  such  that  crystallization
      occurs, centrifuge  to settle the crystals,  and carefully  draw off the
      sample extract with  a  disposable pi pet leaving the excess sulfur  in the K-
      D tube.  Transfer 1.0 mL of the extract to a calibrated centrifuge tube.

            7.1.3 Add approximately 2 g of cleaned  copper  powder  (to the 0.5 mL
      mark) to the centrifuge  tube.  Mix  for at least  1 min on  the mechanical
      shaker.

            7.1.4 Separate  the extract from the  copper by  drawing  off the
      extract with a disposable pipet and  transfer  to a  clean vial.  The volume
      remaining still represents 1.0 mL of extract.

            NOTE: This separation is necessary to prevent further degradation of
                  the pesticides.

      7.2   Removal  of sulfur using mercury

      NOTE: Mercury is a highly toxic metal.  All  operations involving mercury
            should be  performed  in  a  hood.   Prior to  using mercury,  it  is
            recommended that the analyst become acquainted with proper handling
            and cleanup techniques associated with this metal.

            7.2.1 Concentrate  the  sample  extract  to  exactly 1.0  mL  or  other

                                  3660A - 3                         Revision 1
                                                                     July 1992

-------
known  volume.    Perform  concentration  using  the Kuderna-Danish  (K-D)
Technique (Method 3510, Sections 7.10.1 through 7.10.4).

      CAUTION:    When  the  volume  of solvent  is reduced  below  1  ml,
                  semi volatile analytes may be lost.

      7.2.2 Pipet 1.0 ml of  the  extract into a clean concentrator tube or
Teflon-sealed vial.

      7.2.3 Add  one  to three  drops  of  mercury to  the vial  and  seal.
Agitate the contents of the  vial  for 15-30 sec.  Prolonged shaking (2 hr)
may be required.  If so, use a mechanical shaker.

      7.2.4 Separate  the  sample  from the  mercury by  drawing  off  the
extract with a disposable pipet and transfer to a  clean vial.

7.3   Removal of sulfur using TBA sulfite

      7.3.1 Concentrate the  sample extract to  exactly 1.0  ml  or  other
known  volume.    Perform  concentration  using  the Kuderna-Danish  (K-D)
Technique (Method 3510, Sections 7.10.1 through 7.10.4).

      CAUTION:    When  the  volume  of solvent  is reduced  below  1  ml,
                  semivolatile analytes may be lost.

      7.3.2 Transfer 1.0 ml of the extract to a 50 ml clear glass bottle
or vial with a Teflon-lined screw-cap.  Rinse the  concentrator tube with
1 ml of hexane, adding the rinsings to the 50 ml bottle.

      7.3.3 Add 1.0 mL TBA  sulfite  reagent  and  2  ml  2-propanol, cap the
bottle, and shake for  at  least  1  min.   If the sample is colorless or if
the initial  color is unchanged, and if clear  crystals (precipitated sodium
sulfite)  are  observed, sufficient  sodium sulfite is  present.   If the
precipitated  sodium  sulfite  disappears,  add  more  crystalline  sodium
sulfite in  approximately  0.100  g  portions  until a solid residue remains
after repeated shaking.

      7.3.4 Add 5 ml organic free reagent water and shake for at least 1
min.  Allow the sample to stand for 5-10 min.  Transfer the hexane layer
(top) to a concentrator tube and concentrate the extract to approximately
1.0  ml with  the  micro K-D  Technique (Section 7.3.5)  or  the Nitrogen
Slowdown Technique (Section  7.3.6).  Record the actual  volume of the final
extract.

      7.3.5 Micro-Snyder Column Technique

            7.3.5.1     Add another one or two clean boiling  chips to the
      concentrator tube and  attach a two ball micro-Snyder column.  Prewet
      the column  by adding  about 0.5 ml  of hexane  to the top  of the
      column.   Place  the  K-D apparatus in a  hot  water bath so that the
      concentrator tube is  partially  immersed in  the hot water.  Adjust
      the vertical position  of the apparatus and the water temperature, as
      required,  to  complete the concentration in  5-10  minutes.   At the
      proper rate of  distillation the balls of the column will actively

                             3660A - 4                        Revision 1
                                                               July 1992

-------
            chatter,  but the chambers will not flood.  When the apparent volume
            of liquid reaches 0.5 ml, remove the K-D  apparatus  from the water
            bath and  allow  it to drain and cool  for at  least 10 minutes.  Remove
            the Snyder  column  and rinse the flask  and its lower  joints  with
            about 0.2 mL of solvent and add to  the  concentrator tube.   Adjust
            the final volume to approximately  1.0 mL with hexane.

            7.3.6 Nitrogen Slowdown Technique

                  7.3.6.1     Place the concentrator tube in  a warm water bath
            (approximately 35°C) and evaporate  the solvent volume to 1.0-2.0 ml,
            using a  gentle  stream of clean,  dry nitrogen  (filtered through a
            column of activated carbon).

                  CAUTION:     Do not use plasticized tubing between the carbon
                              trap and the sample.

                  7.3.6.2     The internal wall  of the tube must be rinsed down
            several   times  with  the appropriate solvent during  the operation.
            During evaporation,  the solvent level in  the tube must be positioned
            to prevent water from condensing into the  sample (i.e., the solvent
            level should be below  the level of  the  water  bath).   Under normal
            operating conditions, the extract should  not  be  allowed to become
            dry.

                  CAUTION:     When the volume of solvent is reduced below 1 mL,
                              semi volatile analytes  may be lost.

      7.4   Analyze  the cleaned  up  extracts  by gas  chromatography  (see  the
determinative methods, Section 4.3 of this chapter).


8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One for specific  quality  control  procedures and
Method 3600 for cleanup procedures.

      8.2   All  reagents  should  be. checked  prior  to  use  to  verify  that
interferences do not exist.
9.0   METHOD PERFORMANCE

      9.1   Table 1 indicates the effect of using copper and mercury to remove
sulfur on the recovery of certain pesticides.


10.0  REFERENCES

1.    Loy, E.W., private communication.

2.    Goerlitz, D.F. and L.M.  Law, Bulletin for Environmental Contamination and
      Toxicology, 6, 9 (1971).


                                  3660A - 5                         Revision 1
                                                                     July 1992

-------
3.    U.S.  EPA Contract  Laboratory Program,  Statement of  Work  for  Organic
      Analysis, Revision, July 1985.
                                   3660A - 6                         Revision  1
                                                                      July 1992

-------
                                  Table 1.
                  EFFECT OF MERCURY AND COPPER ON PESTICIDES
                                       Percent Recovery3 using:
  Pesticide                         Mercury                 Copper
Aroclor 1254
Lindane
Heptachlor
Aldrin
Heptachlor epoxide
DDE
DDT
BHC
Dieldrin
Endrin
Chi orobenzi late
Malathion
Diazinon
Parathion
Ethion
Trithion
97.10
75.73
39.84
95.52
69.13
92.07
78.78
81.22
79.11
70.83
7.14
0.00
0.00
0.00
0.00
0.00
104.26
94.83
5.39
93.29
96.55
102.91
85.10
98.08
94.90
89.26
0.00
0.00
0.00
0.00
0.00
0.00
a Percent  recoveries  cited are averages  based  on duplicate  analyses  for all
  compounds  other than  for  Aldrin and  BHC.    For  Aldrin,  four  and  three
  determinations were  averaged to  obtain the result for mercury  and  copper,
  respectively.  Recovery of BHC using copper is based on one analysis.
                                  3660A  - 7                 ,        Revision 1
                                                                     July 1992

-------
                               METHOD  3660A
                             SULFUR CLEANUP
                           7 1  1
                        Concentrate
                          sample
                         ex tract
    7  1 2
 Centrifuge
and draw off
   sample
  ex tract
                           712
                         Transfer
                        extract  to
                        centrifuge
                           tube
                      i
             721
          Concentrate
            sample
           extract
          722 Pipet
         extract into
         concentrator
         tube or vial
           7 2 3 Add
           mercury,
           agitate
1      L
   741
Concentrate
  sample
 extract
   732
 Transfer
extract to
centrifuge
   tube.
 733  Add
TBA-sulfite
    and
2-propanol,
 agitate
                                  3660A -  8
                                            Revision 1
                                             July  1992

-------
                           METHOD 3660A
                             continued
                 1^1
  7  1 3 Add
  copper
powder,  mix
   7 2 4
 Separate
sample from
 mercury
    7 1 4
  Separate
ex tract from
   copper
 7 3  3 Add
more  sodium
 sulfite,
  shake
                     7 3 4 Add
                      reagent
                   wa ter, shake,
                    concent rate
                     extract
                  Analyze extract
                 using appropriate
                   determinative
                    procedure
                             3660A  -  9
                                                   Revision  1
                                                    July 1992

-------
                                  METHOD  3665

                      SULFURIC ACID/PERMANGANATE CLEANUP
1.0   SCOPE AND APPLICATION

      1.1   This method is suitable for the rigorous cleanup of sample extracts
prior to analysis  for  polychlorinated biphenyls.   This method  should  be used
whenever elevated  baselines  or overly complex chromatograms  prevent accurate
quantitation of PCBs.  This method cannot  be used to cleanup extracts for other
target  analytes,   as  it  will  destroy  most  organic  chemicals   including  the
pesticides  Aldrin,  Dieldrin,  Endrin,  Endosulfan  (I  and  II),  and  Endosulfan
sulfate.
2.0   SUMMARY OF METHOD

      2.1   An  extract  is  solvent  exchanged  to  hexane,  then  the  hexane  is
sequentially treated with (1)  concentrated  sulfuric acid  and, if necessary, (2)
5% aqueous potassium permanganate.  Appropriate caution must be taken with these
corrosive reagents.

      2.2   Blanks and replicate analysis  samples must  be subjected to the same
cleanup as the samples associated with them.

      2.3   It is important that all  the extracts be exchanged to hexane before
initiating the following treatments.


3.0   INTERFERENCES

      3.1   This technique will  not  destroy chlorinated  benzenes,  chlorinated
naphthalenes (Halowaxes), and a number of chlorinated pesticides.


4.0   APPARATUS

      4.1   Syringe or Class A volumetric pipet, glass;  1.0, 2.0 and 5.0 mL.

      4.2   Vials -1,2 and 10  mL, glass with  Teflon lined  screw caps or crimp
tops.

      4.3   Kuderna-Danish (K-D) apparatus.

            4.3.1 Concentrator tube - 10 mL graduated (Kontes K-570050-1025 or
      equivalent).   A ground  glass  stopper is  used to prevent  evaporation  of
      extracts.

            4.3.2 Evaporation    flask  -  500  mL  (Kontes    K-570001-500    or
      equivalent).   Attach  to  concentrator  tube  with   springs,  clamps,  or
      equivalent.
                                   3665 - 1                         Revision 0
                                                                September 1994

-------
            4.3.3 Snyder  column  - Three  ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.3.4 Snyder  column   -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.3.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).

      4.4   Vortex mixer.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used in  all tests. Unless
otherwise  indicated,  it  is  intended  that all  reagents  shall conform  to  the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where  such  specifications  are  available.  Other grades may  be used,
provided it is first  ascertained that the reagent is  of sufficiently high purity
to permit its use without lessening  the accuracy of the determination.

      5.2   Organic-free reagent  water.   All references to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Sulfuric acid/Water,  H2S04/H20, (1:1, v/v).

      5.4   Hexane, C6H14  - Pesticide grade or equivalent.

      5.5   Potassium permanganate,  KMn04,  5  percent aqueous solution (w/v).


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material to this chapter,  Organic Analytes, Sec.
4.1.


7.0   PROCEDURE

      7.1   Sulfuric acid cleanup

            7.1.1 Using a syringe or  a volumetric pipet, transfer 1.0 or 2.0 mL
      of the hexane extract to a  10  mL vial and, in a fume hood,  carefully add
      5 mL of the 1:1 sulfuric acid/water solution.

            7.1.2 The volume of hexane extract used  depends  on the requirements
      of  the  GC  autosampler  used  by  the laboratory.    If   the  autosampler
      functions reliably with 1 mL of sample volume,  1.0 mL of  extract should be
      used.  If the autosampler requires more than 1  mL of  sample volume, 2.0 mL
      of extract should be used.

            CAUTION:     Make  sure that  there  is no  exothermic reaction  nor
                        evolution of gas prior to proceeding.
                                   3665 - 2                         Revision 0
                                                                September 1994

-------
      7.1.3 Cap the vial tightly and vortex for one minute.   A vortex must
be visible in the vial.

      CAUTION:    Stop the vortexing immediately if the vial leaks, AVOID
                  SKIN CONTACT, SULFURIC ACID BURNS.

      7.1.4 Allow the phases to separate for at least 1 minute.  Examine
the top (hexane) layer;  it  should not be highly colored nor  should it have
a visible emulsion or cloudiness.

      7.1.5 If  a  clean   phase  separation   is  achieved,  proceed  to
Sec. 7.1.8.

      7.1.6 If the hexane  layer is  colored or the emulsion persists for
several minutes, remove the sulfuric acid  layer  from the vial and dispose
of it properly.  Add another 5 ml of the clean 1:1 sulfuric acid/water.

      NOTE: Do not remove any hexane at this  stage of the procedure.

      7.1.7 Vortex the  sample  for  one minute  and allow  the  phases  to
separate.

      7.1.8 Transfer the hexane layer to a clean 10 ml vial.

      7.1.9 Add an additional 1 ml of hexane  to the sulfuric acid layer,
cap and  shake.   This  second  extraction is  done  to ensure quantitative
transfer of the PCBs and Toxaphene.

      7.1.10      Remove the  second hexane layer  and combine  with the
hexane from Sec. 7.1.8.

7.2   Permanganate cleanup

      7.2.1 Add 5  ml of  the 5  percent aqueous  potassium permanganate
solution to the combined hexane fractions  from 7.1.10.

      CAUTION:    Make  sure  that there is  no  exothermic  reaction  nor
                  evolution of gas prior to  proceeding.

      7.2.2 Cap the vial tightly and vortex  for 1 minute.   A vortex must
be visible in the vial.

      CAUTION:    Stop the  vortexing immediately if the vial  leaks.  AVOID
                  SKIN CONTACT,  POTASSIUM  PERMANGANATE BURNS.

      7.2.3 Allow the phases to separate for  at least 1 minute.  Examine
the top (hexane) layer,  it should not be highly colored nor  should it have
a visible emulsion or cloudiness.

      7.2.4 If  a  clean  phase  separation   is   achieved,   proceed  to
Sec.  7.2.7.
                             3665 - 3                         Revision 0
                                                          September 1994

-------
      7.2.5 If the hexane layer  is  colored  or  the  emulsion persists for
several minutes,  remove  the permanganate solution  from the vial  via  a
glass pipette and dispose of it properly.  Add another 5 ml of the clean
aqueous permanganate solution.

      NOTE: Do not remove any hexane at this stage of the procedure.

      7.2.6 Vortex the sample and allow the phases to separate.

      7.2.7 Transfer the hexane layer to a clean 10 ml vial.

      7.2.8 Add an additional  1  ml  of  hexane  to the permanganate layer,
cap the vial  securely and shake.  This second extraction  is done to ensure
quantitative transfer of the PCBs and Toxaphene.

      7.2.9 Remove the second  hexane layer and combine  with  the hexane
from Sec.  7.2.7.

7.3   Final preparation

      7.3.1 Reduce  the volume  of  the  combined  hexane  layers   to  the
original  volume   (1   or  2  ml)   using  the   Kuderna-Danish  Technique
(Sec. 7.3.1.1).

            7.3.1.1     Add one or  two  clean  boiling  chips to the flask
      and attach a three  ball Snyder column.  Prewet the Snyder column by
      adding about 1 ml of hexane to the top of the  column.   Place  the K-D
      apparatus on a hot water bath (15-20°C  above  the  boiling point of
      the solvent) so  that  the  concentrator  tube is partially immersed in
      the hot water and the entire lower rounded surface of the flask is
      bathed  with hot vapor.    Adjust the  vertical  position   of  the
      apparatus and the water  temperature,  as  required,  to complete the
      concentration in 10-20 minutes.   At the proper rate of distillation
      the balls of the  column will actively chatter,  but  the chambers will
      not  flood.   When the  apparent volume  of liquid  reaches  1-2 mL,
      remove the K-D apparatus  from the water bath and allow it to drain
      and cool for at  least 10 minutes.

            7.3.1.2     Remove the Snyder column and rinse the flask and
      its lower joints into the concentrator tube with 1-2 mL of  hexane.
      The extract  may  be further concentrated  by  using  either the micro
      Snyder column technique (Sec.  7.3.2) or nitrogen blowdown technique
      (Sec. 7.3.3).

      7.3.2 Micro  Snyder Column Technique

            7.3.2.1     Add another one or  two  clean boiling chips to the
      concentrator tube and  attach a two ball micro Snyder column.  Prewet
      the  column  by adding  about  0.5  mL  of  hexane  to the  top  of the
      column.   Place  the  K-D apparatus  in  a hot water  bath so that the
      concentrator tube is  partially  immersed  in  the  hot water.   Adjust
      the vertical position  of  the apparatus and the water temperature, as
      required, to complete  the concentration  in 5-10  minutes.   At the


                             3665 - 4                         Revision 0
                                                          September 1994

-------
            proper rate of  distillation  the  balls  of the column will actively
            chatter, but the chambers will  not flood.  When the apparent volume
            of liquid  reaches  0.5  ml,  remove the K-D apparatus from the water
            bath and allow it to drain  and cool for at least  10 minutes.  Remove
            the  Snyder column and rinse the flask and  its  lower  joints with
            about 0.2 ml of  hexane  and  add to the concentrator tube.  Adjust the
            final volume to  1.0-2.0 ml, as required, with hexane.

            7.3.3 Nitrogen Slowdown Technique

                  7.3.3.1      Place the concentrator tube in  a warm water bath
            (approximately  35°C)  and  evaporate  the  solvent  volume  to  the
            required  level   using  a   gentle  stream  of  clean,  dry  nitrogen
            (filtered  through  a  column of activated carbon).

                  CAUTION:     Do not use plasticized tubing between the carbon
                               trap and the sample.

                  7.3.3.2      The internal  wall of the tube must be rinsed down
            several times with the appropriate  solvent  during  the operation.
            During evaporation, the solvent level in the tube  must be positioned
            to prevent water from condensing  into the  sample  (i.e., the solvent
            level should be  below  the  level  of the water bath).   Under normal
            operating  conditions,  the  extract  should  not be  allowed to become
            dry.

            7.3.4 Remove  any  remaining  organochlorine  pesticides  from  the
      extracts using Florisil Column Cleanup  (Method 3620) or  Silica Gel Cleanup
      (Method 3630).

            7.3.5 The  extracts  obtained may now  be  analyzed for  the  target
      analytes using the appropriate  organic  technique(s)  (see Sec. 4.3 of this
      Chapter).  If analysis of  the extract will not be performed immediately,
      stopper the concentrator tube and store in a refrigerator.   If the extract
      will be stored longer  than 2 days, it should be transferred  to a vial with
      a Teflon lined screw cap or crimp top,  and labeled appropriately.
8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality control procedures.


9.0   METHOD PERFORMANCE

      9.1   No performance data are currently available.


10.0  REFERENCES

      None required.
                                   3665 - 5                         Revision 0
                                                                September 1994

-------
                                METHOD 3665
                SULFURIC  ACID/PERMANGANATE  CLEANUP
    Start
7.1.1  Carefully
combine hexane
   with 1:1
  H2S04/H20
   solution.
    7.1.2
 Transfer the
  appropriate
  volume to
     vial.
 7.1.3 - 7.1.4
  Cap. vortex
   and allow
    phase
  separation.
    7.1.8
   Transfer
 hexane layer
 to clean vial.
                         7.1.6 Remove
                          and dispose
                        H2SO4 solution,
                        add clean HZS04
                           solution.
7.1.7 Cap,
vortex, and
allow phase
separation.
   7.1.9 Add
   hexane to
  H2SO4 layer,
 cap and shake.
    7.1.10
 Combine two
 hexane layers.
                                                     7.2.1 Add
                                                      KMn04
                                                      solution.
                                                     7.2.2 - 7.2.3
                                                     Cap, vortex,
                                                      and allow
                                                       phase
                                                     separation.
/ 7.2.4 Is N.
/ phase \ No w
V. separation )~~
N. clean? /
[Yes
7..
Tra
hexan
to cle
1
2.7
isfer
an vial.
f
7.2.8 Add
hexane to
KMn04 layer.
cap and shake.
>
r
7.2.9 Combine
two hexane
layers.


	

w

7.2.5 Remove
and dispose
KMnQ4 aotution,
add clean KMnO4
solution.
V
7.2.6 Cap
vortex and
separation.

7.3.1 - 7.3.3
Reduce volumn
using K-D
and/or nitrogen
btowdown tech.
V

7.3.4 Use
Method 3620 or
Method 3630 to
further remove
contaminants.
                                                                                   7.3.5 Stopper
                                                                                       and
                                                                                     refrigerate
                                                                                     for further
                                                                                     analysis.
                                                                                        Stop
                                    3665  -  6
                                                        Revision 0
                                                  September  1994

-------
4.3  DETERMINATION OF ORGANIC ANALYTES
     4.3.1 GAS CHROMATOGRAPHIC METHODS

         The following methods are included in this section:
         Method 8000A:
         Method 8010B:

         Method 8011:

         Method 8015A:

         Method 8020A:
         Method 8021A:
         Method 8030A:
         Method 8031:
         Method 8032:
         Method 8040A:
         Method 8060:
         Method 8061:

         Method 8070:
         Method 8080A:

         Method 8081:

         Method 8090:
         Method 8100:
         Method 8110:
         Method 8120A:
         Method 8121:

         Method 8140:
         Method 8141A:

         Method 8150B:
         Method 8151:
Gas Chromatography
Halogenated    Volatile    Organics    by    Gas
Chromatography
1,2-Dibromoethane and l,2-Dibromo-3-chloropropane
by Microextraction and Gas Chromatography
Nonhalogenated   Volatile    Organics    by   Gas
Chromatography
Aromatic Volatile Organics by Gas Chromatography
Halogenated Volatiles  by  Gas  Chromatography Using
Photoionization  and  Electrolytic  Conductivity
Detectors in Series: Capillary Column Technique
Acrolein and Acrylonitrile by Gas Chromatography
Acrylonitrile by Gas Chromatography
Acrylamide by Gas Chromatography
Phenols by Gas Chromatography
Phthalate Esters
Phthalate Esters by Capillary Gas Chromatography
with Electron Capture  Detection (GC/ECD)
Nitrosamines by Gas Chromatography
Organochlorine  Pesticides  and  Polychlorinated
Biphenyls by Gas Chromatography
Organochlorine Pesticides and PCBs as Aroclors by
Gas Chromatography:  Capillary Column Technique
Nitroaromatics and Cyclic Ketones
Polynuclear Aromatic Hydrocarbons
Haloethers by Gas Chromatography
Chlorinated Hydrocarbons  by Gas Chromatography
Chlorinated Hydrocarbons  by  Gas Chromatography:
Capillary Column Technique
Organophosphorus Pesticides
Organophosphorus Compounds by Gas Chromatography:
Capillary Column Technique
Chlorinated Herbicides by Gas Chromatography
Chlorinated Herbicides by GC  Using Methylation or
Pentafluorobenzylation Derivatization:  Capillary
Column Technique
                                 FOUR - 10
                                      Revision 2
                                  September 1994

-------
                                 METHOD 8000A

                              GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION
      1.1   Gas  chromatography is  a quantitative  technique  useful  for  the
analysis  of organic  compounds  capable of  being  volatilized without  being
decomposed or  chemically  rearranged.   Gas chromatography  (GC),  also  known as
vapor phase chromatography (VPC), has two subcategories distinguished by: gas-
solid chromatography  (GSC),  and  gas-liquid  chromatography  (GLC)  or gas-liquid
partition chromatography  (GLPC).   This  last group  is  the  most commonly used,
distinguished by type of column adsorbent or packing.

      1.2   The chromatographic methods are recommended for use only by, or under
the close supervision of,  experienced residue analysts.


2.0   SUMMARY OF METHOD

      2.1   Each organic analytical method that follows provides a recommended
technique  for  extraction, cleanup,  and occasionally,  derivatization  of  the
samples to be analyzed.  Before the prepared sample is introduced into the GC,
a procedure for standardization must be followed to determine the recovery and
the  limits of  detection   for  the  analytes of  interest.    Following  sample
introduction into the GC,  analysis proceeds with a comparison of sample values
with standard values.   Quantitative analysis  is achieved through integration of
peak area or measurement of peak height.


3.0   INTERFERENCES

      3.1   Contamination by  carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.  To reduce carryover,  the
sample syringe or purging device must be rinsed out between samples with water
or solvent. Whenever  an unusually concentrated sample  is encountered, it should
be followed by  an  analysis of  a solvent blank or of  water to  check for cross
contamination.   For volatile samples containing large amounts of water-soluble
materials,  suspended  solids,  high  boiling  compounds or high  organohalide
concentrations, it may be necessary  to  wash  out  the syringe or purging device
with a detergent solution, rinse it with distilled water,  and then dry it in a
105°C oven between  analyses.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph - Analytical system complete with gas chromatograph
suitable  for   on-column  injections  and all  required accessories,  including
detectors, column supplies,  recorder, gases, and  syringes.   A  data system for
measuring peak height and/or peak areas is recommended.

      4.2   Gas chromatographic columns  - See the  specific determinative method.
Other packed or capillary  (open-tubular) columns may be used if the requirements

                                   8000A -  1                         Revision 1
                                                                     July 1992

-------
of Section 8.6 are met.


5.0   REAGENTS

      5.1   See the specific determinative method for the reagents needed.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Section 4.1.


7.0   PROCEDURE

      7.1   Extraction - Adhere to those procedures specified in the referring
determinative method.

      7.2   Cleanup and separation - Adhere to those procedures specified in the
referring determinative method.

      7.3   The recommended gas chromatographic columns and operating conditions
for the instrument are specified  in the referring determinative method.

      7.4   Calibration

            7.4.1 Establish gas chromatographic operating parameters equivalent
      to those indicated in Section 7.0 of the determinative method of interest.
      Prepare   calibration  standards  using  the   procedures   indicated  in
      Section 5.0  of  the  determinative method  of  interest.    Calibrate the
      chromatographic  system  using   either  the  external  standard  technique
      (Section  7.4.2)  or the internal standard technique (Section 7.4.3).

            7.4.2 External  standard calibration procedure

                  7.4.2.1     For each analyte of interest, prepare calibration
            standards  at a  minimum of five concentrations by adding volumes of
            one or more  stock  standards  to  a volumetric flask and diluting to
            volume with an  appropriate solvent.  One of the external standards
            should be  at a  concentration near, but above, the method detection
            limit.  The other concentrations should correspond to the expected
            range of concentrations found in real samples or should define the
            working range  of the  detector.

                  7.4.2.2     Inject  each   calibration  standard  using  the
            technique  that will  be used to introduce the actual samples  into the
            gas chromatograph  (e.g.  2-5  pi injections, purge-and-trap, etc.).
            Tabulate peak  height  or  area responses against the mass injected.
            The results  can be used  to  prepare a calibration  curve  for each
            analyte.   Alternatively,  for samples that are introduced into the
            gas chromatograph using a syringe, the ratio of the  response to the
            amount  injected,  defined as  the calibration factor  (CF),  can be
            calculated for  each analyte at each  standard concentration.  If the

                                   8000A - 2                        Revision  1
                                                                     July 1992

-------
percent relative standard deviation (%RSD)  of the calibration factor
is less  than 20%  over  the working  range,  linearity  through  the
origin can be  assumed,  and the average calibration  factor  can be
used in place of a calibration curve.

      Calibration factor -

      *     For multiresponse  pesticides/PCBs, use the total area of
            all peaks used for quantitation.

      7.4.2.3     The  working  calibration  curve  or  calibration
factor must be verified  on  each  working day  by the injection of one
or more  calibration  standards.   The frequency of  verification is
dependent on  the detector.  Detectors, such  as the electron capture
detector,  that  operate   in  the   sub-nanogram   range  are  more
susceptible to changes in detector response  caused by GC column and
sample   effects.     Therefore,   more  frequent   verification   of
calibration  is  necessary.   The flame ionization detector is much
less  sensitive  and requires  less frequent  verification.   If  the
response for  any analyte varies  from  the predicted response by more
than  +  15%,  a  new calibration  curve  must be  prepared  for that
analyte.   For  methods 8010,   8020, and 8030,  see Table  3 in each
method for calibration and quality control acceptance criteria.

                           R!  -  R2
      Percent Difference = 	  x 100
                             RI

where:

      R,    =     Calibration Factor from first analysis.

      R2    =     Calibration Factor from succeeding analyses.

7.4.3 Internal standard calibration  procedure

      7.4.3.1     To use this  approach, the  analyst must  select one
or more  internal standards that are  similar in analytical behavior
to the compounds of interest.  The analyst must further demonstrate
that  the measurement  of the   internal standard is  not affected by
method  or  matrix  interferences.    Due  to  these  limitations,  no
internal standard applicable to all  samples can be suggested.

      7.4.3.2     Prepare calibration standards at a minimum of five
concentrations for each  analyte  of interest by  adding  volumes  of one
or more stock standards to  a volumetric flask.  To each calibration
standard,  add  a  known  constant  amount  of one  or  more internal
standards and dilute to volume with an appropriate  solvent.  One of
the  standards  should  be at a concentration near, but above,  the
method detection limit.  The other concentrations should correspond
to the  expected range of concentrations   found in  real samples or
should define the working range of the detector.
                       8000A -  3                         Revision 1
                                                         July 1992

-------
            7.4.3.3     Inject each calibration standard using  the same
      introduction technique that will be applied to  the actual  samples
      (e.g.  2 to 5 /xL injection, purge-and-trap,  etc.).  Tabulate the peak
      height or area responses  against the concentration of each compound
      and internal standard.   Calculate  response factors  (RF)  for each
      compound  as follows:

            RF     -     (AsCis)/(AisCs)

      where:

            As     =     Response for the  analyte to  be measured.

            Ais    =     Response for the  internal  standard.

            Cjs    =     Concentration of  the internal  standard,  /xg/L.

            Cs     =     Concentration of the  analyte to  be  measured,
                        M9/L.

            If the RF  value over the working  range is constant  (< 20%
      RSD),  the RF can  be assumed to  be invariant, and the average RF can
      be used for  calculations.  Alternatively,  the results  can be used to
      plot a calibration curve of response ratios, As/Ajs versus RF.

            7.4.3.4     The  working  calibration  curve  or  RF  must  be
      verified on  each working day  by  the measurement  of one  or more
      calibration standards.  The frequency of verification is dependent
      on the detector.   Detectors,  such as the electron capture detector,
      that  operate  in   the  sub-nanogram  range  are  more susceptible to
      changes  in detector response caused by GC column  and sample effects.
      Therefore, more frequent verification of calibration  is necessary.
      The flame ionization  detector  is much less  sensitive and requires
      less frequent verification.   If the response for any analyte varies
      from the predicted response  by more  than  +  15%,  a  new calibration
      curve must be prepared for that compound.   For methods 8010, 8020,
      and 8030, see Table  3 in each method for calibration  and quality
      control  acceptance criteria.

7.5   Retention time windows

      7.5.1 Before establishing windows, make sure the GC  system is within
optimum  operating conditions.   Make three  injections of  all  single
component  standard  mixtures  and  multiresponse  products   (i.e.  PCBs)
throughout the course  of a  72  hour period.   Serial  injections over less
than a 72 hour  period result in retention  time windows  that  are too tight.

      7.5.2 Calculate the standard  deviation of  the  three retention times
(use any function of retention time; including  absolute retention time, or
relative  retention  time)   for each  single  component  standard.    For
multiresponse  products, choose  one major  peak  from the  envelope and
calculate the  standard deviation  of the three  retention times  for that
peak.    The peak chosen  should  be fairly  immune  to  losses   due to
degradation and weathering  in  samples.

                             8000A  -  4                        Revision  1
                                                               July  1992

-------
            7.5.2.1     Plus or minus three times the standard deviation
      of the  retention times for each standard will  be used to define the
      retention time window; however, the experience of the analyst should
      weigh  heavily  in  the  interpretation  of  chromatograms.    For
      multiresponse  analytes  (i.e.  PCBs),  the  analyst  should  use  the
      retention  time  window,   but  should  primarily  rely  on  pattern
      recognition.

            7.5.2.2     In those cases where the  standard deviation for a
      particular  standard  is  zero,  the  laboratory must  substitute  the
      standard deviation of a close eluting,  similar compound to develop
      a valid retention time window.

      7.5.3 The laboratory must  calculate retention  time windows for each
standard on each GC column  and whenever a new  GC column is installed.  The
data must be retained by the laboratory.

7.6   Gas chromatographic analysis

      7.6.1 Introduction of organic compounds into the gas chromatograph
varies depending on the volatility of the compound.   Volatile organics are
primarily introduced by purge-and-trap  (Method  5030).  However, there are
limited  applications  (in  Method   5030)  where  direct   injection   is
acceptable.   Use of Method  3810 or 3820  as  a  screening  technique  for
volatile organic  analysis  may  be valuable with  some sample  matrices to
prevent overloading  and  contamination  of the  GC  systems.   Semivolatile
organics are introduced by direct injection.

      7.6.2 The appropriate detector(s) is given  in the specific method.

      7.6.3 Samples  are  analyzed  in a  set referred  to as  an analysis
sequence.   The  sequence  begins with instrument  calibration  followed by
sample   extracts   interspersed  with   multi-concentration   calibration
standards.   The  sequence ends when the set of  samples has been injected or
when qualitative and/or quantitative QC criteria  are exceeded.

      7.6.4 Direct Injection -  Inject 2-5 juL of the sample extract using
the solvent flush technique, if the extract is  manually injected.  Smaller
volumes (1.0 /xL) can be injected, and the solvent flush technique is not
required, if automatic devices  are employed.   Record the volume injected
to the nearest 0.05 p.1 and the resulting peak size in area units or peak
height.

      7.6.5 If the responses exceed the linear range of the system, dilute
the extract and reanalyze.   It  is recommended that extracts be diluted so
that all peaks  are on  scale.   Overlapping peaks  are not  always evident
when  peaks are  off  scale.    Computer  reproduction  of  chromatograms,
manipulated to ensure all  peaks  are  on  scale  over a 100-fold range,  are
acceptable  if linearity  is demonstrated.   Peak  height  measurements  are
recommended over peak area integration when overlapping peaks cause errors
in area integration.

      7.6.6 If   peak  detection  is  prevented   by  the  presence   of
interferences, further cleanup  is required.

                            8000A  -  5                         Revision 1
                                                               July 1992

-------
            7.6.7 Examples of chromatograms for the compounds  of  interest  are
      frequently available in the referring analytical  method.

            7.6.8 Calibrate  the  system  immediately  prior  to  conducting  any
      analyses (see Section  7.4).   A mid-concentration standard must  also  be
      injected at  intervals  specified  in  the method  and at  the  end of  the
      analysis  sequence.    The  calibration  factor  for  each  analyte  to  be
      quantitated,  must not exceed a 15% difference when compared to the initial
      standard of  the analysis  sequence.   When  this  criterion is  exceeded,
      inspect  the  GC  system  to determine  the  cause and  perform  whatever
      maintenance  is  necessary  (see Section  7.7)  before  recalibrating  and
      proceeding with  sample  analysis.  All  samples that were injected after the
      standard exceeding  the criterion  must be reinjected to  avoid  errors  in
      quantitation, if  the  initial  analysis  indicated  the presence of  the
      specific target  analytes  that  exceeded the criterion.

            7.6.9 Establish daily retention time windows for each analyte.  Use
      the retention time for  each analyte from Section 7.6.8 as  the midpoint of
      the window  for  that day.   The daily  retention  time  window equals  the
      midpoint ± three times the standard deviation  determined  in  Section 7.5.

                  7.6.9.1     Tentative identification of an  analyte occurs when
            a peak from a sample extract falls within  the  daily retention time
            window. Normally, confirmation is required: on a second GC column,
            by  GC/MS  if  concentration   permits,  or  by   other   recognized
            confirmation techniques.   Confirmation may not be necessary if the
            composition  of the  sample  matrix  is well  established   by  prior
            analyses.

                  7.6.9.2     Validation of GC system  qualitative  performance:
            Use  the  mid-concentration  standards  interspersed  throughout  the
            analysis sequence (Section  7.6.8) to  evaluate this criterion.   If
            any of the standards  fall outside their daily retention time window,
            the system is  out of control.   Determine  the  cause of  the problem
            and correct it (see  Section  7.7).  All  samples  that were injected
            after  the  standard  exceeding  the  criteria must be reinjected  to
            avoid false negatives and possibly false positives.

      7.7   Suggested  chromatography system maintenance - Corrective measures may
require any one or more of the following remedial  actions.

            7.7.1 Packed columns  - For  instruments  with  injection  port traps,
      replace the demister trap,  clean,  and deactivate the glass injection port
      insert or  replace  with a  cleaned  and deactivated insert.   Inspect  the
      injection end of the column and  remove any foreign material (broken glass
      from the rim of the  column or  pieces of septa).   Replace the glass wool
      with fresh deactivated glass wool.   Also,  it  may be necessary  to remove
      the first few millimeters  of the packing material if any discoloration is
      noted, also  swab  out the  inside  walls  of  the column if any  residue is
      noted.  If these procedures fail to eliminate the degradation problem, it
      may be  necessary to deactivate the  metal  injector body (described  in
      Section 7.7.3) and/or repack/replace the column.
                                   8000A -  6                         Revision 1
                                                                     July 1992

-------
      7.7.2 Capillary columns - Clean and deactivate the glass injection
port insert or replace with a cleaned and deactivated insert.  Break off
the first few inches, up  to  one  foot,  of the injection port side of the
column.   Remove  the  column  and  solvent  backflush  according   to  the
manufacturer's instructions.   If these  procedures  fail  to eliminate the
degradation problem, it may be necessary to deactivate the metal injector
body and/or replace the column.

      7.7.3 Metal  injector  body  -  Turn off  the  oven  and  remove  the
analytical column when  the oven  has  cooled.   Remove the glass injection
port insert (instruments  with off-column injection or Grob).  Lower the
injection port temperature  to room temperature.   Inspect the injection
port and remove any noticeable foreign material.

            7.7.3.1     Place a beaker  beneath the injector port inside
      the GC oven.  Using a wash  bottle,  serially rinse the entire inside
      of the  injector  port with acetone  and  then  toluene;  catching the
      rinsate in the beaker.

            7.7.3.2     Prepare a solution of deactivating agent (Sylon-CT
      or equivalent) following manufacturer's directions.  After all metal
      surfaces inside the injector body have been thoroughly coated with
      the deactivation  solution,  serially rinse the  injector body with
      toluene, methanol, acetone, and nexane.  Reassemble the injector and
      replace the GC column.

7.8   Calculations

      7.8.1 External standard calibration  -  The  concentration  of each
analyte  in  the sample  may  be determined  by calculating the  amount of
standard purged or injected,  from the peak response, using the calibration
curve  or the  calibration  factor  determined  in  Section  7.4.2.   The
concentration of a specific analyte is calculated as follows:

      Aqueous samples

      Concentration (/ig/L) »  [(Ax)(A)(Vt)(D)]/[(As)(V1)(V8)]

where:

      Ax    =     Response for the  analyte in the sample,  units may be in
                  area counts or peak height.

      A     =     Amount of standard injected or purged, ng.

      As    =     Response for the external  standard,  units same as for
                  V

      V,    =     Volume  of  extract  injected,  piL.   For purge-and-trap
                  analysis, V1 is not applicable and therefore =  1.

      D     =     Dilution factor,  if dilution was  made on  the sample
                  prior to  analysis.    If no  dilution was made,  D  = 1,
                  dimensionless.

                             8000A  -  7                          Revision 1
                                                               July 1992

-------
      Vt    -     Volume  of  total  extract,   /uL.    For  purge-and-trap
                  analysis, Vt is not applicable and therefore = 1.

      V8    =     Volume of sample extracted or purged, ml.


Nonaqueous samples

      Concentration (Mg/kg) -  I(AJl)(A)(Vt)(D)]/[(AB)(Vl)(W)]

where:

      W     =     Weight of sample extracted or purged, g.  The wet  weight
                  or dry weight may be used, depending upon  the  specific
                  applications of the data.

      Ax, A8,  A, Vt, D,  and  V,  have the  same  definition  as for aqueous
samples when a solid sample is purged (e.g., low  concentration soil)  for
volatile  organic analysis  or  for semivolatile  organic  and pesticide
extracts.   When the nonaqueous  sample  is  extracted  for  purge and  trap
analysis, V, «  volume of methanol extract added to reagent water for purge
and trap analysis.

      7.8.2 Internal standard calibration - For each analyte of interest,
the concentration of that analyte in the sample is calculated as  follows:

      Aqueous samples

      Concentration (Mg/L) =  [(Ax)(Cis)(D)]/[(A1s)(RF)(Vs)]

where:

      Ax    •     Response of the analyte being measured,  units may be in
                  area counts or peak height.

      Cjs   =     Amount of internal standard  added to extract or  volume
                  purged, ng.

      D     =     Dilution factor,  if  a dilution was made on  the  sample
                  prior  to  analysis.    If no  dilution was  made, D  = 1,
                  dimensionless.

      Ajs   =     Response of the  internal  standard,  units  same  as Ax.

      RF    =     Response factor  for  analyte, as  determined  in Section
                  7.4.3.3.

      Vs    =     Volume of water  extracted or purged, ml.

      Nonaqueous samples

      Concentration (/tg/kg)  = [(As)(Cis)(D)]/[(Ais)(RF)(W8)]
                             8000A - 8                         Revision 1
                                                                July 1992

-------
      where:

            Ws    =     Weight of sample extracted, g.  Either a dry weight or
                        wet weight  may be  used,  depending upon  the specific
                        application of the data.

            As,  Cjs,  D, Ajs, and  RF  have the  same definition as  for aqueous
      samples.


8.0   QUALITY CONTROL

      8.1   Each laboratory that  uses  these methods is required  to  operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory capability and an ongoing analysis of
spiked samples  to  evaluate and document quality data.  The  laboratory should
maintain records to document the  quality of the data  generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics of the method.
When results of sample  spikes  indicate atypical  method performance,  a quality
control check standard should  be analyzed to confirm that the measurements were
performed in an in-control mode of operation.

      8.2   Before  processing  any  samples, the  analyst   should  demonstrate,
through the analysis of a  reagent  blank, that interferences from the analytical
system, glassware, and reagents are under control.   Each time a set of samples
is extracted or there  is  a change in  reagents,  an organic-free  reagent water
blank  should   be   processed   as  a   safeguard  against   chronic  laboratory
contamination.  The blank  samples should be carried through  all  stages of the
sample preparation and measurement steps.

      8.3   For each analytical batch (up to  20 samples), a reagent blank, matrix
spike, and duplicate or matrix spike duplicate should be analyzed (the frequency
of the spikes may be different  for  different monitoring programs).   The blank and
spiked samples should be carried through all  stages of the sample preparation and
measurement steps.

      8.4   The experience of  the  analyst  performing gas chromatography  is
invaluable to the  success  of the methods.  Each day that analysis is performed,
the  daily  calibration  sample  should  be  evaluated  to  determine  if  the
chromatographic system is  operating properly.   Questions  that should be asked
are: Do  the peaks  look normal?;  Is the response  obtained comparable  to the
response  from  previous calibrations?    Careful  examination  of  the  standard
chromatogram can  indicate  whether the column  is  still good, the  injector is
leaking,  the injector septum needs replacing, etc.   If any changes are made to
the system (e.g. column changed), recalibration of the  system should take place.

      8.5   Required instrument QC

            8.5.1  Step 7.4 requires that the %RSD vary by  < 20% when comparing
      calibration  factors  to  determine if  a five  point calibration  curve is
      linear.
                                  8000A  - 9                         Revision 1
                                                                     July 1992

-------
            8.5.2 Section 7.4 sets a limit of +  15% difference when comparing
      daily  response  of a  given analyte  versus the  initial  response.   For
      Methods 8010, 8020, and 8030,  follow the guidance on limits specified in
      Section 7.4.3.4.   If the limit is exceeded,  a new  standard curve should be
      prepared  unless  instrument maintenance  corrects the  problem for  that
      particular analyte.

            8.5.3 Step 7.5 requires the establishment of retention time windows.

            8.5.4 Section 7.6.8 sets  a  limit of +  15% difference when comparing
      the response from the continuing calibration standard of a given analyte
      versus any succeeding standards analyzed during an analysis sequence.

            8.5.5 Step  7.6.9.2  requires that all  succeeding standards  in an
      analysis  sequence  should  fall  within  the  daily retention  time  window
      established by the first standard of the sequence.

      8.6   To  establish the  ability  to generate acceptable  accuracy  and
precision, the analyst should perform the following  operations.

            8.6.1 A quality control  (QC)  check sample  concentrate  is required
      containing each analyte of interest.  The QC check sample concentrate may
      be  prepared  from  pure standard  materials,  or  purchased as  certified
      solutions.  If  prepared by the laboratory, the  QC  check sample concentrate
      should be  made  using  stock standards prepared independently from those
      used for calibration.

                  8.6.1.1     The  concentration  of  the  QC   check  sample
            concentrate   is  highly   dependent   upon   the   analytes   being
            investigated.  Therefore, refer to Method 3500, Section 8.0 for the
            required concentration of the QC check sample concentrate.

            8.6.2 Preparation of QC check samples

                  8.6.2.1     Volatile organic analytes (Methods 8010, 8020, and
            8030) - The QC check sample is prepared  by  adding 200 p.1 of the QC
            check sample concentrate (Step 8.6.1) to 100 ml of water.

                  8.6.2.2     Semivolatile organic analytes (Methods 8040, 8060,
            8070, 8080,  8090,  8100,  8110, and 8120) -  The  QC  check sample is
            prepared by adding 1.0 ml of the QC check sample concentrate (Step
            8.6.1) to each of four 1-L aliquots of water.

            8.6.3 Four aliquots of the well-mixed QC check sample are analyzed
      by the same procedures used to analyze actual samples (Section 7.0 of each
      of the methods).   For  volatile  organics, the preparation/analysis process
      is purge-and-trap/gas chromatography.  For semivolatile organics, the QC
      check samples should undergo solvent extraction (see Method 3500) prior to
      chromatographic analysis.

            8.6.4 Calculate the  average recovery  (x) in M9/L> and the standard
      deviation of the recovery (s)  in  M9/L,  for  each analyte of interest using
      the four results.
                                  8000A - 10                        Revision 1
                                                                     July 1992

-------
            8.6.5 For  each  analyte  compare  s  and x  with the  corresponding
      acceptance criteria for precision and accuracy, respectively, given the QC
      Acceptance Criteria Table at the end of each  of the determinative methods.
      If s and x for all  analytes of  interest meet the acceptance criteria, the
      system performance is  acceptable and analysis of  actual samples can_begin.
      If any individual s exceeds the precision  limit or any individual x falls
      outside  the  range   for accuracy,  then  the   system   performance  is
      unacceptable for that analyte.

            NOTE: The  large number  of  analytes in  each  of the  QC  Acceptance
                  Criteria Tables present a substantial probability that one or
                  more will fail at  least one of  the  acceptance  criteria when
                  all analytes of a given method are determined.

            8.6.6 When one or more  of the analytes tested fail  at least one of
      the  acceptance  criteria, the  analyst  should proceed according  to  Step
      8.6.6.1 or 8.6.6.2.

                  8.6.6.1     Locate and correct the source of the problem and
            repeat  the  test   for  all   analytes  of   interest  beginning  with
            Step 8.6.2.

                  8.6.6.2     Beginning with Step  8.6.2,  repeat  the  test only
            for those analytes  that failed to meet criteria.  Repeated failure,
            however, will confirm a general problem with the measurement system.
            If this occurs, locate  and  correct  the source of  the  problem and
            repeat  the  test  for  all  compounds  of  interest  beginning  with
            Step 8.6.2.

      8.7   The laboratory should,  on an  ongoing basis, analyze a reagent blank
and a matrix spiked duplicate for each  analytical  batch (up to a  maximum of 20
samples/batch) to assess accuracy.  For soil and waste samples where detectable
amounts of organics are present, replicate samples may be appropriate in place
of spiked duplicates.   For laboratories analyzing  one to ten samples per month,
at least one spiked sample per month is required.

            8.7.1 The  concentration  of  the  spike in  the sample  should  be
      determined as follows:

                  8.7.1.1     If, as  in compliance  monitoring, the concentration
            of a  specific analyte  in  the sample  is  being checked  against  a
            regulatory concentration  limit,  the  spike  should be at that limit,
            or 1 to 5  times  higher than the background concentration determined
            in Step 8.7.2, whichever concentration would  be larger.

                  8.7.1.2     If the concentration of a  specific  analyte  in  a
            water sample is not being checked against  a limit  specific to that
            analyte, the  spike  should  be at  the same concentration  as  the QC
            reference  sample  (Step   8.6.2)  or  1  to  5 times  higher than  the
            background  concentration  determined   in   Step  8.7.2,   whichever
            concentration would be larger.  For  other matrices, the recommended
            spiking concentration  is  20 times the  EQL.


                                 8000A - 11                         Revision  1
                                                                     July 1992

-------
            8.7.1.3     For semivolatile organics,  it may not be possible
      to determine the background  concentration  levels  prior to spiking
      (e.g. maximum holding times will be exceeded).  If  this  is the case,
      the spike concentration should be (1) the regulatory concentration
      limit, if any; or,  if none (2) the larger of either 5 times higher
      than  the  expected background  concentration or  the QC  reference
      sample  concentration  (Step   8.6.2).    For  other  matrices,  the
      recommended spiking concentration is 20 times the  EQL.

      8.7.2 Analyze  one  unspiked   and   one  spiked  sample  aliquot  to
determine percent recovery of each of the spiked compounds.

            8.7.2.1     Volatile  organics  -  Analyze   one  5-mL  sample
      aliquot  to determine  the background  concentration  (B)  of  each
      analyte.    If  necessary,  prepare  a  new  QC  reference  sample
      concentrate   (Step 8.6.1)    appropriate    for    the   background
      concentration in the  sample.   Spike a second  5-mL  sample aliquot
      with 10 fj.1 of the QC  reference sample concentrate  and analyze it to
      determine  the  concentration  after  spiking  (A)  of  each  analyte.
      Calculate each percent recovery (p) as 100(A - B)%/T,  where T is the
      known true value of the spike.

            8.7.2.2     Semivolatile organics - Analyze one  sample aliquot
      (extract of  1-L  sample) to determine  the  background concentration
      (B)  of  each analyte.   If necessary, prepare  a  new QC  reference
      sample  concentrate  (Step  8.6.1)   appropriate  for the  background
      concentration in the  sample.  Spike a second 1-L sample aliquot with
      1.0 mL  of  the QC  reference  sample  concentrate and  analyze  it to
      determine  the  concentration  after  spiking  (A)  of  each  analyte.
      Calculate each percent recovery (p) as 100(A - B)%/T,  where T is the
      known true value of the spike.

      8.7.3 Compare the percent recovery (p) for each analyte in a water
sample  with  the corresponding  criteria  presented in the  QC Acceptance
Criteria Table  found  at the end of each of the determinative  methods.
These acceptance  criteria  were calculated  to  include  an  allowance  for
error in  measurement of both  the  background and  spike concentrations,
assuming a spike to background ratio  of 5:1.  This error  will be accounted
for to the extent that the  analyst's  spike to background ratio approaches
5:1.   If  spiking  was performed at  a concentration  lower than  the QC
reference sample concentration  (Step  8.6.2), the analyst  should use either
the  QC  acceptance  criteria presented  in  the   Tables,  or  optional  QC
acceptance criteria calculated for the specific  spike concentration.   To
calculate  optional  acceptance  criteria for the  recovery  of  an  analyte:
(1)  Calculate accuracy  (x') using  the  equation found  in  the  Method
Accuracy and  Precision as  a Function of Concentration Table  (appears at
the   end   of  each   determinative  method),   substituting   the  spike
concentration  (T)  for  C;  (2) calculate  overall_precision (S')  using the
equation in the same Table,  substituting x' for x; (3) calculate the range
for recovery at the spike concentration as (lOOx'/T)  +  2.44(100S'/T)%.

      8.7.4 If  any  individual  p falls outside the designated  range for
recovery,  that analyte  has failed  the  acceptance  criteria.    A check
standard  containing each  analyte   that  failed  the  criteria should be

                            8000A  -  12                        Revision  1
                                                               July  1992

-------
      analyzed as described in Step 8.8.

      8.8   If any analyte in a water sample fails the acceptance criteria for
recovery in Step 8.7, a QC reference standard containing each analyte that failed
should be prepared and analyzed.

      NOTE: The frequency for the required analysis of a QC reference standard
            will  depend upon the number of analytes being simultaneously tested,
            the complexity  of the sample  matrix,  and the  performance  of the
            laboratory.  If the entire list of analytes given in  a method should
            be measured  in  the sample  in  Step 8.7,  the  probability  that the
            analysis of a QC check standard will be required is high.   In this
            case, the QC  check  standard  should be routinely analyzed  with the
            spiked sample.

            8.8.1 Preparation of the QC  check  sample  - For volatile organics,
      add 10 fj.1 of the  QC check  sample  concentrate (Step  8.6.1  or 8.7.2) to 5
      ml of water. For  semivolatile organics, add 1.0  ml of the  QC check sample
      concentrate (Step 8.6.1 or  8.7.2)  to 1 L of water.  The  QC  check sample
      needs only  to  contain  the analytes that failed  criteria  in  the test in
      Step  8.7.    Prepare  the  QC  check  sample  for  analysis  following  the
      guidelines given in Method 3500 (e.g. purge-and-trap, extraction, etc.).

            8.8.2 Analyze the QC  check  sample to  determine the concentration
      measured (A) of  each  analyte.   Calculate  each  percent recovery  (ps) as
      100(A/T)%,  where T  is the true value of the standard concentration.

            8.8.3 Compare the percent recovery  (ps) for each  analyte  with the
      corresponding  QC  acceptance  criteria found in the  appropriate  Table in
      each of the methods.  Only analytes that  failed the  test in Step 8.7 need
      to be compared with these criteria.  If the recovery of any such analyte
      falls outside  the designated range,  the  laboratory  performance  for that
      analyte  is  judged  to  be  out of  control,  and  the  problem should  be
      immediately identified and corrected.  The  result for  that analyte in the
      unspiked  sample   is suspect  and   may  not  be   reported   for  regulatory
      compliance purposes.

      8.9   As part  of the QC program for  the  laboratory, method  accuracy for
each matrix studied  should be  assessed and  records should  be maintained.  After
the analysis of five spiked samples  (of  the  same  matrix type)  as  in Step 8.7,
calculate the  average  percent recovery  (p) and the standard deviation  of the
percent recovery  (s ).   Express the  accuracy assessment as a percent  recovery
interval from p - 2s to  p  +  2s .   If  p = 90% and s   = 10%,  for  example, the
accuracy interval  is expressed as 70-110%.  Update the accuracy assessment for
each  analyte  on  a regular  basis  (e.g.  after  each five  to ten  new  accuracy
measurements).

      8.10  Calculate surrogate control  limits  as follows:

            8.10.1      For each sample analyzed, calculate the percent recovery
      of each surrogate in the sample.

            8.10.2      Calculate the average percent recovery (p) and standard
      deviation of the  percent  recovery  (s) for  each of  the  surrogates when

                                  8000A - 13                        Revision 1
                                                                     July 1992

-------
      surrogate data from 25 to 30 samples for each matrix is available.

            8.10.3      For  a  given  matrix,  calculate  the  upper and  lower
      control limit for method performance  for  each  surrogate standard.   This
      should be done as follows:

            Upper Control  Limit (UCL) = p + 3s
            Lower Control  Limit (LCL) = p - 3s

            8.10.4      For  aqueous  and   soil   matrices,   these  laboratory
      established surrogate control  limits  should,  if  applicable,  be  compared
      with the  control  limits in  Tables A  and B of  Methods 8240 and  8270,
      respectively.    The  limits  given  in these methods  are multi-laboratory
      performance based limits  for  soil  and  aqueous samples, and therefore, the
      single-laboratory limits established  in Step 8.10.3  should  fall  within
      those given in Tables A and B for these matrices.

            8.10.5      If  recovery  is  not  within  limits,  the  following  is
      required.

            •     Check  to  be  sure  there  are  no  errors   in  calculations,
                  surrogate  solutions  and  internal  standards.    Also,  check
                  instrument performance.

            •     Recalculate the data  and/or reanalyze the  extract  if any of
                  the above checks reveal a problem.

            •     Reextract and reanalyze the  sample  if none  of the above are a
                  problem or flag the data as "estimated concentration."

            8.10.6      At a minimum, each  laboratory  should update  surrogate
      recovery limits on a matrix-by-matrix basis, annually.

      8.11  It  is  recommended that  the  laboratory  adopt  additional  quality
assurance practices for use with  this method.   The specific practices that are
most productive depend upon the needs of the  laboratory and the nature of the
samples.   Field duplicates  may  be  analyzed  to  assess the precision of the
environmental measurements. When doubt exists over the  identification of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar column,  specific  element  detector, or mass  spectrometer  should be
used.   Whenever possible,  the laboratory  should analyze  standard  reference
materials and participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1   The  method  detection  limit  (MDL)   is  defined  as  the  minimum
concentration  of a  substance that  can  be  measured  and  reported  with  99%
confidence that the value is above zero.  The MDL concentrations listed in the
referring analytical methods were  obtained  using  water.   Similar results were
achieved using representative wastewaters.  The MDL actually achieved in a given
analysis will vary depending on instrument sensitivity and matrix effects.
                                  8000A - 14                        Revision 1
                                                                     July 1992

-------
      9.2   Refer to the determinative method  for  specific method performance
information.
10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final  Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

2.    U.S.  EPA Contract  Laboratory Program,  Statement of  Work  for Organic
      Analysis, July 1985, Revision.
                                  8000A - 15                        Revision 1
                                                                     July 1992

-------
                      METHOD 8000A
                   GAS CHROMATOGRAPHY
Start
X^ s


7 1 Refer to
determinative
procedure for
extraction
procedure
recommendation

Internal Standard External Standard
7431 Select
internal standards
having behavior
similar to
compounds of
interest
1
7 2 Refer to
determinative
procedure for
cleanup and
preparation
procedure
recommendation*

7432 Prepare
calibration
s tandards
1
741 Establish
chroma togr a phi c
condi tions


/ • /

7433 Inject
cal ibra tion
standards ,
calculate RF
1
7434 Verify
working calibration
curve or RF each
day
S 1 4 l >v
< 	 f Calibration ) 	 »
^v technique? /
~l 5 Calculate
windows

7 4,2.1 Prepare
cal ibration
standards for each
compound of
interest
i
7422 Inject
t d d
1 fa t ' ^ '*
or calculate
calibration factor
1
7423 Verify
working cal ibration
curve each day


                         8000A - 16
Revision 1
 July 1992

-------
                                      METHOD  8000A
                                        continued
                                          Semivolatiles
     7 6 1 If
    necessary.
  screen sample*
  by Method 3810
     or 3820
  761 Introduce
 compounds into  CC
 by purge-and-trap
or direct infection
  (Method 5030)
761 Introduce
compounds  into
 CC by direct
  injection
 764  Inject
 samples  using
 sol vent  flush
  technique,
record  volume
                                                     765 Dilute
                                                      extract  and
                                                      reanalyze
                                                       7  6  6  Do
                                                        fur ther
                                                       cleanup

  6 8 Calibrate
    system
  immedia tely
   prior  to
   analyses
7 6 9 Establish
daily retention
 time windows
   for each
   analyte
                         7  7  Perform
                       chroma tography
                           sys tern
                       maintenance, if
                           needed
                        7  8  Calculate
                      concentration of
                     each  analyte, using
                     appropriate formula
                     for matrix and type
                        of standard
                                                                                Stop
                                        8000A  -  17
                                Revision  1
                                  July  1992

-------
                                 METHOD 8010B

              HALOGENATED  VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8010  is  used  to  determine  the  concentration of  various
volatile  halogenated  organic  compounds.    The  following  compounds  can  be
determined by this method:
Appropriate Technique
Compound Name
Allyl chloride
Benzyl chloride
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl ) ether
Bromoacetone
Bromobenzene
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethanol
2-Chloroethyl vinyl ether
Chloroform
1-Chlorohexane
Chl oromethane
Chloromethyl methyl ether
Chloroprene
4-Chlorotoluene
Di bromochl oromethane
1 , 2-Di bromo-3-chl oropropane
Dibromomethane
1, 2-Di chlorobenzene
1, 3 -Di chlorobenzene
1, 4 -Di chlorobenzene
l,4-Dichloro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
tr ans-1, 2-Di chl oroethene
Dichloromethane
1,2-Dichloropropane
l,3-Dichloro-2-propanol
ci s- 1 ,3-Di chl oropropene
trans-l,3-Dichloropropene
Epichlorhydrin
CAS No.a
107-05-1
100-44-7
111-91-1
39638-32-9
598-31-2
108-86-1
75-27-4
75-25-2
74-83-9
56-23-5
108-90-7
75-00-3
107-07-03
110-75-8
67-66-3
544-10-5
74-87-3
107-30-2
126-99-8
106-43-4
124-48-1
96-12-8
74-95-3
95-50-1
541-73-1
106-46-7
764-41-0
75-71-8
75-34-3
107-06-2
75-35-4
156-60-5
75-09-2
78-87-5
96-23-1
10061-01-5
10061-02-6
106-89-8
Purge-and-Trap
b
PP
PP
b
PP
b
b
b
b
b
b
b
PP
b
b
pc
b
PP
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
PP
b
b
PP
Direct
Injection
b
b
pc
b
b
b
b
b
b
b
b
b
b
b
b
pc
b
pc
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
                                  8010B  - 1
    Revision 2
September 1994

-------
                                                     Appropriate Technique
                                                                      Direct
Compound Name                        CAS No.a       Purge-and-Trap     Injection
Ethylene dibromide
Methyl iodide
1,1,2,2-Tetrachloroethane
1,1,1 , 2-Tetrachl oroethane
Tetrachl oroethene
1,1,1 -Tr i chl oroethane
1,1, 2 -Trichl oroethane
Trichloroethene
Tr i chl orof 1 uoromethane
1,2,3-Trichloropropane
Vinyl Chloride
106-93-4
74-88-4
79-34-5
630-20-6
127-18-4
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
75-01-4
b
PP
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
a  Chemical Abstract Services Registry Number
b  Adequate response using this technique
pp Poor purging efficiency, resulting in high EQLs
pc Poor chromatographic performance.

      1.2   Table 1 indicates compounds that may be  analyzed by this method and
lists the method detection limit for  each compound in  organic-free reagent water.
Table 2 lists the estimated quantitation limit for other matrices.


2.0   SUMMARY OF METHOD

      2.1   Method  8010   provides  gas  chromatographic  conditions  for  the
detection of halogenated volatile organic compounds.  Samples can be introduced
into the GC using direct injection or purge-and-trap (Method 5030).  Ground water
samples must be analyzed using Method 5030.   A temperature program is  used in the
gas chromatograph to separate the organic compounds.  Detection is achieved by
a electrolytic conductivity detector (HECD).

      2.2   The method provides an optional  gas  chromatographic column that may
be helpful  in resolving the analytes  from co-eluting non-target  compounds and for
analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Methods  5030 and  8000.

      3.2   Samples  can  be  contaminated  by diffusion  of volatile  organics
(particularly chl orofluorocarbons and  methylene chloride)  through  the sample
container  septum  during  shipment and storage.   A  trip blank  prepared  from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as  a check on such contamination.
                                   8010B -  2                         Revision 2
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS

      4.1    Gas chromatograph

             4.1.1  Gas chromatograph  -  analytical  system  complete  with  gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and all required accessories,  including detector, analytical
      columns, recorder, gases, and syringes.  A  data  system  for measuring peak
      heights and/or  peak  areas is  recommended.

             4.1.2  Columns

                   4.1.2.1    Column 1 - 8 ft  x  0.1  in.  ID stainless steel  or
             glass  column  packed with 1% SP-1000  on  Carbopack-B  60/80 mesh  or
             equivalent.

                   4.1.2.2    Column 2 - 6 ft  x  0.1  in.  ID stainless steel  or
             glass  column  packed with chemically  bonded  n-octane  on Porasil-C
             100/120 mesh  (Durapak)  or equivalent.

             4.1.3  Detector -  Electrolytic  conductivity  (HECD).

      4.2    Sample  introduction  apparatus,   refer  to   Method  5030  for  the
appropriate equipment for  sample introduction purposes.

      4.3    Syringes, 5 ml Luerlok  glass hypodermic and a 5 ml, gas-tight with
shutoff valve.

      4.4    Volumetric  flask,  Class  A,  Appropriate  sizes  with  ground  glass
stoppers.

      4.5    Microsyringe,  10 and 25 juL with a 0.006 in. ID needle (Hamilton 702N
or equivalent) and a 100 //L.

      4.6    Analytical balance  - 0.0001 g.


5.0   REAGENTS

      5.1    Reagent  grade  chemicals shall  be  used  in  all  tests.    Unless
otherwise  indicated,  it  is  intended  that  all reagents  shall conform  to  the
specifications of the  Committee on  Analytical Reagents of the American Chemical
Society, where such specifications are  available.  Other grades  may  be used,
provided it is first ascertained that the reagent  is of sufficiently high purity
to permit its use without  lessening the accuracy  of the determination.

      5.2    Organic-free reagent water.  All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3    Methanol, CH3OH.  Pesticide  quality or equivalent.  Store away from
other solvents.
                                   8010B  -  3                         Revision 2
                                                                September 1994

-------
      5.4   Stock standards - Stock solutions may be prepared  from pure standard
materials or  purchased as  certified  solutions.   Prepare stock  standards  in
methanol using assayed  liquids or gases,  as appropriate.  Because of the toxicity
of some  of  the organohalides,  primary dilutions of these  materials  should  be
prepared in a hood.

            5.4.1 Place about  9.8  ml  of methanol in a 10 ml tared ground glass
      stoppered volumetric  flask.   Allow the  flask to stand, unstoppered, for
      about 10 minutes until all  alcohol-wetted  surfaces have dried.  Weigh the
      flask to the nearest  0.0001  g.

            5.4.2 Add  the  assayed  reference material, as  described below.

                  5.4.2.1     Liquids.  Using  a  100 /LtL syringe,  immediately add
            two or more drops of assayed reference material to the flask; then
            reweigh.   The  liquid must fall directly  into  the  alcohol  without
            contacting the  neck of the flask.

                  5.4.2.2     Gases.  To  prepare standards  for any  compounds
            that   boil   below    30°C   (e.g.   bromomethane,    chloroethane,
            chloromethane,   dichlorodifluoromethane,   trichlorofluoromethane,
            vinyl chloride),  fill   a  5  ml valved  gas-tight  syringe with the
            reference  standard  to  the 5.0  ml mark.   Lower the needle  to 5  mm
            above  the  methanol  meniscus.    Slowly  introduce the  reference
            standard  above  the surface  of the  liquid.  The  heavy gas  rapidly
            dissolves  in the methanol.  This may also be accomplished by using
            a  lecture  bottle equipped with  a Hamilton Lecture Bottle Septum
            (#86600).   Attach  Teflon  tubing to  the side-arm relief  valve and
            direct a gentle stream of gas  into  the methanol  meniscus.

            5.4.3 Reweigh,  dilute  to volume,  stopper,  and  then  mix by inverting
      the flask several times.   Calculate  the concentration  in milligrams per
      liter (mg/L) from the net gain in weight.  When compound purity is assayed
      to  be 96%  or  greater,  the  weight  may be  used  without  correction  to
      calculate the concentration of the stock standard.  Commercially prepared
      stock standards may be used at any concentration if they  are certified by
      the manufacturer or by an independent source.

            5.4.4 Transfer the stock standard  solution  into  a  bottle with  a
      Teflon lined screw-cap.  Store,  with minimal  headspace, at -10°C  to  -20°C
      and protect from light.

            5.4.5 Prepare  fresh  stock standards for gases weekly or sooner if
      comparison with  check standards indicates  a problem.  Reactive compounds
      such as 2-chloroethyl vinyl ether may need to  be  prepared  more frequently.
      All  other  standards  must  be replaced after  six  months.   Both  gas and
      liquid  standards must be  monitored  closely by comparison to the initial
      calibration curve  and by comparison to  QC check standards.   It may be
      necessary to replace the  standards more frequently if either check exceeds
      a  20% drift.

            5.4.6 Optionally calibration using  a certified gaseous mixture can
      be  accomplished  daily utilizing commercially  available  gaseous analyte


                                   8010B  -  4                         Revision  2
                                                                September 1994

-------
      mixture  of bromomethane,  chloromethane,  chloroethane,  vinyl  chloride,
      dichlorodifluoromethane  and  trichlorofluoromethane  in  nitrogen.  These
      mixtures  of documented  quality are  stable  for  as  long as  six months
      without  refrigeration.  (VOA-CYL III,  RESTEK  Corporation, Cat.  #20194 or
      equivalent).

      5.5    Secondary  dilution  standards.  Using   stock  standard  solutions,
prepare  secondary  dilution standards  in  methanol,  as  needed,  containing  the
compounds of interest,  either singly or mixed together.  The secondary dilution
standards should be prepared at concentrations such  that  the aqueous calibration
standards prepared in Sec. 5.6 will  bracket  the working  range of the analytical
system.  Secondary dilution standards should be stored with minimal headspace for
volatiles  and  should  be checked  frequently  for  signs  of  degradation  or
evaporation, especially just prior to preparing calibration standards from them.

      5.6    Calibration   standards.      Prepare   calibration   standards   in
organic-free reagent water from the secondary dilution of the stock standards,
at a minimum of five concentrations.  One of the concentrations should be at a
concentration  near,  but  above,  the  method detection  limit.   The  remaining
concentrations  should  correspond  to the  expected range of  the concentrations
found  in  real   samples or should define  the  working range  of the GC.   Each
standard should contain each  analyte for  detection by  this method (e.g. some or
all of  the  compounds listed  in Table 1 may be  included).  In  order to prepare
accurate aqueous standard  solutions, the following precautions must be observed.

             5.6.1 Do  not  inject more  than  20  /uL of  alcoholic standards into
      100 mL of water.

             5.6.2 Use  a   25  /^L  Hamilton  702N  microsyringe  or  equivalent
      (variations  in   needle  geometry will  adversely  affect  the ability  to
      deliver reproducible volumes  of methanolic standards into water).

             5.6.3 Rapidly inject   the   alcoholic   standard  into  the  filled
      volumetric flask.  Remove the  needle as fast as  possible after injection.

             5.6.4 Mix  aqueous standards by inverting the flask three times only.

             5.6.5 Fill the sample syringe from the standard solution contained
      in the expanded  area of the flask  (do not use any solution  contained in
      the neck  of the  flask).

             5.6.6 Never  use  pipets  to  dilute  or transfer samples  or aqueous
      standards.

             5.6.7 Aqueous standards are not stable and should  be discarded after
      one hour, unless properly  sealed  and  stored.   The aqueous standards  can
      be stored up to  24  hours, if  held  in  sealed vials with zero headspace.

      5.7    Internal standards (if  internal standard  calibration is used) - To
use this approach, the analyst must  select one or more internal standards that
are similar  in  analytical   behavior  to  the compounds of  interest.   The analyst
must further demonstrate  that the measurement of the  internal  standard is  not
affected by method or  matrix  interferences.   Because  of these  limitations,  no


                                  8010B  - 5                        Revision 2
                                                                September 1994

-------
internal  standard  can be  suggested  that is applicable  to all  samples.   The
compounds recommended  for use as  surrogate  spikes  (Sec.  5.8)  have  been used
successfully as internal  standards, because  of  their generally  unique retention
times.

             5.7.1  Prepare  calibration  standards   at   a  minimum   of  five
      concentrations  for  each analyte of interest as described  in Sec.  5.6.

             5.7.2  Prepare a spiking solution  containing  each  of the internal
      standards using the procedures described  in  Sees.  5.4  and 5.5.   It is
      recommended  that  the  secondary  dilution  standard  be  prepared at  a
      concentration  of  15  ng/pL  of each  internal  standard  compound.   The
      addition of  10 juL  of  this  standard  to 5.0 ml of  sample or calibration
      standard would  be equivalent to 30 /ug/L.

             5.7.3  Analyze each calibration  standard according to  Sec.  7.0,
      adding  10  juL  of  internal   standard  spiking  solution  directly  to  the
      syringe.

      5.8    Surrogate  standards   -   The  analyst   should  monitor  both  the
performance  of the analytical system and  the  effectiveness of  the  method in
dealing  with  each  sample  matrix  by   spiking  each  sample,   standard,  and
organic-free reagent water blank with surrogate halocarbons.  A combination of
bromochloromethane, bromochlorobenzene and  bromofluorobenzene  is  recommended to
encompass the  range  of temperature  program  used  in this  method.   From stock
standard solutions prepared as in  Sec.  5.4,  add a volume to give  750 /j,g  of each
surrogate to 45 ml  of organic-free  reagent water contained in a  50 ml volumetric
flask, mix,  and dilute to volume for  a concentration of 15 ng//zL.  Add 10 /xL of
this surrogate spiking solution directly into the 5 ml syringe with every sample
and reference standard analyzed.  If the internal  standard calibration procedure
is used, the surrogate compounds may  be added directly to the internal standard
spiking solution  (Sec. 5.7.2).


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1    See the  introductory  material  to  this  Chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1    Volatile  compounds are  introduced into the gas chromatograph using
either direct injection or purge-and-trap (Method 5030).  Method 5030 may be used
directly on  ground  water samples or  low-concentration  contaminated soils and
sediments.   For medium-concentration soils  or sediments, methanolic extraction,
as described in Method 5030,  may  be  necessary prior to purge-and-trap analysis.

      7.2    Gas  chromatographic  conditions  (Recommended)

             7.2.1  Column 1:

             Helium  flow  rate = 40 mL/min


                                  8010B -  6                        Revision 2
                                                                September 1994

-------
            Temperature  program:
                   Initial  temperature  =   45°C, hold for 3 minutes
                   Program  =              45°C to 220°C  at 8°C/nnn
                   Final  temperature  =     220°C, hold for 15 minutes.

            7.2.2  Column 2:

            Helium  flow  rate  =  40 mL/min
            Temperature  program:
                   Initial  temperature  =   50°C, hold for 3 minutes
                   Program  =              50°C to 170°C  at 6°C/min
                   Final  temperature  =     170°C, hold for 4 minutes.

      7.3   Calibration.  The procedure for internal or external calibration may
be used.  Refer to Method 8000 for a description of each of these procedures. Use
Table 1 and Table 2 for guidance on selecting the lowest point on the calibration
curve.

            7.3.1  Calibration must take place  using the same sample introduction
      method that will be  used  to analyze actual samples  (see Sec.  7.4.1).

      7.4   Gas chromatographic analysis

            7.4.1  Introduce volatile compounds into the  gas  chromatograph using
      either Method 5030 (purge-and-trap) or  the direct injection method (see
      Sec. 7.4.1.1).   If the  internal  standard calibration technique is used,
      add 10 juL of  internal standard to the sample  prior  to purging.

                   7.4.1.1     In very limited applications (e.g. aqueous process
            wastes) direct injection  of  the  sample onto  the  GC  column  with a
            10 jiiL  syringe  may be appropriate.   The detection limit is very high
            (approximately 10,000 M9/L)  therefore,  it is only permitted where
            concentrations in excess of 10,000 p.g/1 are expected or for water-
            soluble compounds that do not purge.  The  system must be calibrated
            by direct  injection (bypassing the  purge-and-trap device).

            7.4.2  Method 8000 provides instructions on  the analysis sequence,
      appropriate  dilutions,  establishing daily retention time  windows,  and
      identification criteria.  Include a mid-concentration  standard after each
      group of 10 samples  in  the analysis sequence.

            7.4.3  Table  1  summarizes  the  estimated retention times  on the two
      columns for  a number of organic  compounds analyzable using this method.
      An example of the  separation achieved by Column 1 is  shown in Figure 1.

            7.4.4  Record the  sample volume purged or injected  and the resulting
      peak sizes (in area  units or peak heights).

            7.4.5  Refer  to  Method   8000   for  guidance  on  calculation  of
      concentration.

            7.4.6  If analytical interferences are suspected, or for the purpose
      of confirmation, analysis using the second GC column  is recommended.


                                  8010B  - 7                         Revision 2
                                                                September 1994

-------
             7.4.7  If the response for  a peak is off-scale,  i.e.,  beyond the
      calibration range of the standards, prepare a  dilution of the sample with
      organic-free  reagent water.  The  dilution  must be performed on a second
      aliquot of the sample which has been properly sealed and stored prior to
      use.


8.0   QUALITY CONTROL

      8.1    Refer  to Chapter One for specific quality  control  procedures and
Method 8000 for gas chromatographic procedures.  Quality control  to ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2    Quality control  required to  validate  the GC  system  operation is
found in Method 8000.

             8.2.1  The  quality control check  sample  concentrate  (Method 8000)
      should contain each analyte of interest at a concentration of 10 mg/L in
      methanol.

             8.2.2  Table  3 indicates the calibration  and  QC  acceptance criteria,
      for water  samples,  for this method.  Table 4  gives  method  accuracy and
      precision  as  functions  of concentration,  for water  samples,  for  the
      analytes  of  interest.    The  contents of both  Tables should be  used to
      evaluate a laboratory's  ability to perform and generate acceptable data
      by this method.

      8.3    Calculate  surrogate  standard recovery on all  samples,  blanks, and
spikes.  Determine if recovery is  within limits (limits established by performing
QC procedure outlined in Method 8000).

             8.3.1  If recovery  is not within limits,  the following is required:

                   •    Check  to  be  sure   that   there   are  no  errors  in
             calculations,  surrogate  solutions and  internal standards.   Also,
             check  instrument performance.

                   •     Recalculate  the data and/or  re-analyze the  sample if
             any  of  the above checks reveal a  problem.

                   •    Re-extract  and  re-analyze  the  sample  if  none  of the
             above are  a  problem or flag  the data as "estimated concentration".


9.0   METHOD PERFORMANCE

      9.1    This method was tested by 20 laboratories using  organic-free reagent
water, drinking  water,  surface water, and three industrial wastewaters spiked at
six concentrations  over  the range 8.0-500  jug/L.   Single  operator precision,
overall  precision,  and  method accuracy were found to  be directly related to the
concentration of the analyte,  and essentially  independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 4.
                                   8010B -  8                         Revision 2
                                                                September 1994

-------
      9.2    The accuracy and precision obtained will be determined by the sample
matrix,  sample introduction technique, and by the calibration procedure used.

      9.3    The method detection limits reported in Table 1 were generated under
optimum analytical conditions by an Agency contractor  (Ref. 6) as guidance, and
may not be readily achievable by all laboratories at all times.


10.0  REFERENCES

1.    Bellar, T.A.; Lichtenberg, J.J. vL.  Amer. Water Works Assoc. 1974, 66(12),
      pp. 739-744.

2.    Bellar,  T.A.;  Lichtenberg,  J.J.,  Semi-Automated Headspace  Analysis  of
      Drinking  Waters and  Industrial  Waters for Purgeable  Volatile Organic
      Compounds, Measurement of Organic Pollutants in  Water and Wastewater; Van
      Hall,  Ed.; ASTM STP 686, pp  108-129, 1979.

3.    "Development and Application of Test Procedures for Specific Organic Toxic
      Substances  in  Wastewaters: Category 11 -  Purgeables and  Category  12  -
      Acrolein,  Acrylonitrile,   and  Dichlorodifluoromethane";  report  for  EPA
      Contract 68-03-2635.

4.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act: Final  Rule and Interim
      Final  Rule and Proposed Rule", October 26,  1984.

5.    "EPA Method  Validation Study  23,  Method  601  (Purgeable  Halocarbons)";
      report for EPA Contract 68-03-2856.

6.    Gebhart, J.E., S.V. Lucas, S.J. Naber,  A.M. Berry, T.H.  Danison and H.M.
      Burkholder, "Validation of SW-846  Methods  8010,  8015,  and 8020"; Report
      for  EPA  Contract  68-03-1760,  Work Assignment  2-15;    US   EPA,  EMSL-
      Cincinnati, 1987.
                                   8010B  -  9                         Revision 2
                                                                September 1994

-------
                       TABLE  1.
CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
           FOR HALOGENATED VOLATILE ORGANICS
Compound
ATlyl chloride*^
Benzyl chloride*'0
Bis(2-chloroethoxy)methane*
Bis(2-chloroisopropyl ) ether*
Bromobenzene
Bromodichl oromethane
Bromoform*
Bromomethane*
Carbon tetrachloride*
Chloroacetaldehyde*
Chlorobenzene*
Chloroethane
Chloroform*
1-Chlorohexane
2-Chloroethyl^vinyl ether*
Chloromethane*
Chloromethyl methyl ether*
4-Chlorotoluene
Di bromochl oromethane
l,2-Dibromo-3-chloropropane*
Dibromomethane*
1 , 2-Di chl orobenzene*
1,3-Dichlorobenzene*
1, 4 -Di chl orobenzene*
l,4-Dichloro-2-butene*
Di chl orodi f 1 uoromethane*'d
1,1-Dichloroethane*
1, 2-Di chloroethane*
1, 1-Dichloroethene*
trans -1, 2-Di chl proethene*
Di chl oromethane*
1 , 2-Di chl oropropane*
trans-l,3-Dichloropropene*
Ethylene dibromide
1,1,2 , 2 -Tetrachl oroethane*
1,1, 1,2 -Tetrachl oroethane*
Tetrachloroethene*
1 , 1 , 1-Tri chl oroethane^
1 , 1 , 2-Tri chl oroethane*
CAS
Registry
Number
107-05-1
100-44-7
111-91-1
39638-32-9
108-86-1
75-27-4
75-25-2
74-83-9
56-23-5
107-20-0
108-90-7
75-00-3
67-66-3
544-10-5
110-75-8
74-87-3
107-30-2
106-43-4
124-48-1
96-12-8
74-95-3
95-50-1
541-73-1
106-46-7
764-41-0
75-71-8
75-34-3
107-06-2
75-35-4
156-60-5
75-09-2
78-87-5
10061-02-5
106-93-4
79-34-5
630-20-6
127-18-4
71-55-6
79-00-5
Retention Time
(minutes)
Column 1 Column 2
10.17
30.29
38.60
34.79
29.05
15.44
21.12
2.90
14.58
(b)
25.49
5.18
12.62
26.26
19.23
1.40
8.88
34.46
18.22
28.09
13.83
37.96
36.88
38.64
23.45
3.68
11.21
13.14
10.04
11.97
7.56
16.69
16.976
19.59
23.12
21.10
23.05
14.48
18.27
(b)
(b)
(b)
(b)
(b)
14.62
19.17
7.05
11.07
(b)
18.83
8.68
12.08
(b)
(b)
5.28
(b)
(b)
16.62
(b)
14.92
23.52
22.43
22.33
(b)
(b)
12.57
15.35
7.72
9.38
10.12
16.62
16.60
(b)
(b)
21.70
14.97
13.10
18.07
Method
Detection
Limit3
(M9/L)
(b)
(b)
(b)
(b)
(b)
0.002
0.020
0.030
0.003
(b)
0.001
0.008
0.002
(b)
0.130
0.010
(b)
(b)
(b)
0.030
(b)
(b)
(b)
(b)
(b)
(b)
0.002
0.002
0.003
0.002
(b)
(b)
0.340
(b)
0.010
(b)
0.001
0.003
0.007
                       8010B  -  10
    Revision 2
September 1994

-------
                                   TABLE 1.
                                   Continued

Compound
CAS
Registry
Number
Retention Time
(minutes)
Column 1 Column 2
Method
Detection
Limit3
(M9/L)
Trichloroethene
Trichlorofluoromethane*
1,2,3-Trichloropropane*
Vinyl Chloride*
                              79-01-6
                              75-69-4
                              96-18-4
                              75-01-4
         17.40
          9.26
         22.95
          3.25
13.12
 (b)
 (b)
 5.28
0.001
 (b)
 (b)
0.006
a =
b =
* =
c =
d =

e =
Using purge-and-trap method (Method 5030). See Sec. 9.3.
Not determined
Appendix VIII compounds
Demonstrated very erratic results
See  Sec.  4.10.2  of Method  5030
material
Estimated retention time
when tested by purge-and-trap
for  guidance on  selection  of  trapping
                                   TABLE 2.
             DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                             FOR VARIOUS MATRICES3
               Matrix
                                               Factor
               Ground water                             10
               Low-concentration soil                    10
               Water miscible liquid waste             500
               High-concentration soil  and sludge     1250
               Non-water miscible waste               1250
               EQL = [Method detection limit (see Table 1)] X [Factor found in
               this table].  For non-aqueous samples,  the  factor  is  on  a wet-
               weight basis.  Sample EQLs are highly matrix-dependent.  The EQLs
               listed herein are provided  for  guidance and may not  always be
               achievable.
                                  8010B - 11
                                                                Revision 2
                                                            September 1994

-------
                            TABLE 3.
             CALIBRATION AND QC ACCEPTANCE CRITERIA3
Range Limit
for Q for S
Analyte
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethylvinyl ether
Chloroform
Chl oromethane
Di bromochl oromethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 , 4-Di chl orobenzene
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-l,2-Dichloroethene
Dichloromethane
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans- 1,3-Di chl oropropene
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
1,1,1-Tri chl oroethane
1, 1, 2 -Tri chl oroethane
Trichloroethene
Trichlorofluoromethane
Vinyl chloride
(M9/L) (M9/L)
15.2-24.8
14.7-25.3
11.7-28.3
13.7-26.3
14.4-25.6
15.4-24.6
12.0-28.0
15.0-25.0
11.9-28.1
13.1-26.9
14.0-26.0
9.9-30.1
13.9-26.1
16.8-23.2
14.3-25.7
12.6-27.4
12.8-27.2
15.5-24.5
14.8-25.2
12.8-27.2
12.8-27.2
9.8-30.2
14.0-26.0
14.2-25.8
15.7-24.3
15.4-24.6
13.3-26.7
13.7-26.3
4.3
4.7
7.6
5.6
5.0
4.4
8.3
4.5
7.4
6.3
5.5
9.1
5.5
3.2
5.2
6.6
6.4
4.0
5.2
7.3
7.3
9.2
5.4
4.9
3.9
4.2
6.0
5.7
Q = Concentration measured in QC check sampl
Range
for x
(M9/L)
10.7-32.0
5.0-29.3
3.4-24.5
11.8-25.3
10.2-27.4
11.3-25.2
4.5-35.5
12.4-24.0
D-34.9
7.9-35.1
1.7-38.9
6.2-32.6
11.5-25.5
11.2-24.6
13.0-26.5
10.2-27.3
11.4-27.1
7.0-27.6
10.1-29.9
6.2-33.8
6.2-33.8
6.6-31.8
8.1-29.6
10.8-24.8
9.6-25.4
9.2-26.6
7.4-28.1
8.2-29.9
e, in /ig/L.
S = Standard deviation of four recovery measurements, in
x = Average recovery
P, Ps = Percent recovery
D = Detected; result
3 r ** 4 + r\ v< •! •» -P^/-.m /in rco D->
Range
P, Ps
(%)
42-172
13-159
D-144
43-143
38-150
46-137
14-186
49-133
D-193
24-191
D-208
7-187
42-143
47-132
51-147
28-167
38-155
25-162
44-156
22-178
22-178
8-184
26-162
41-138
39-136
35-146
21-156
28-163

M9/L.
for four recovery measurements, in /^g/L.
measured.
must be greater
v+ T3C fnv Ma+kn



than zero.
A CHI on
A utnva /*a1/~ii1
a + a/H accnminn
a QC check sample concentration of 20 M9/L.
                            8010B - 12
    Revision 2
September 1994

-------
                                TABLE 4.
       METHOD  ACCURACY  AND PRECISION  AS  FUNCTIONS  OF  CONCENTRATION3
Analyte
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethyl vinyl etherb
Chloroform
Chl oromethane
Di bromochl oromethane
1, 2 -Di chlorobenzene
1,3-Dichlorobenzene
1, 4 -Di chlorobenzene
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans-l,2-Dichloroethene
Di chl oromethane
l,2-Dichloropropaneb
cis-l,3-Dichloropropeneb
trans- 1,3-Di chl oropropeneb
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
1,1,1-Tri chl oroethane
1 , 1 , 2-Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
Vinyl chloride
Accuracy, as
recovery, x'
(M9/L)
1.12C-1.02
0.96C-2.05
0.76C-1.27
0.98C-1.04
l.OOC-1.23
0.99C-1.53
l.OOC
0.93C-0.39
0.77C+0.18
0.94C+2.72
0.93C+1.70
0.95C+0.43
0.93C-0.09
0.95C-1.08
1.04C-1.06
0.98C-0.87
0.97C-0.16
0.91C-0.93
l.OOC
l.OOC
l.OOC
0.95C+0.19
0.94C+0.06
0.90C-0.16
0.86C+0.30
0.87C+0.48
0.89C-0.07
0.97C-0.36
Single analyst
precision, s '
(M9/U
0.11X+0.04
0.12X+0.58
0.28X+0.27
0.15X+0.38
0.15X-0.02
0.14X-0.13
0.20X
0.13X+0.15
0.28X-0.31
0.11X+1.10
0.20X+0.97
0.14X+2.33
0.15X+0.29
0.08X+0.17
0.11X+0.70
0.21X-0.23
0.11X+1.46
0.11X+0.33
0.13X
0.18X
0.18X
0.14X+2.41
0.14X+0.38
0.15X+0.04
0.13X-0.14
0.13X-0.03
0.15X+0.67
0.13X+0.65
Overall
precision,
S' (jug/L)
0.20X+1.00
0.21X+2.41
0.36X+0.94
0.20X+0.39
0.18X+1.21
0.17X+0.63
0.35X
0.19X-0.02
0.52X+1.31
0.24X+1.68
0.13X+6.13
0.26X+2.34
0.20X+0.41
0.14X+0.94
0.15X+0.94
0.29X-0.04
0.17X+1.46
0.21X+1.43
0.23X
0.32X
0.32X
0.23X+2.79
0.18X+2.21
0.20X+0.37
0.19X+0.67
0.23X+0.30
0.26X+0.91
0.27X+0.40
x' =  Expected recovery for one or more measurements of a sample containing
      a concentration of C, in /ug/L.
sr'=  Expected  single analyst  standard  deviation  of  measurements  at  an
      average concentration of x, in ng/l.
S' =  Expected  interlaboratory  standard  deviation  of  measurements  at  an
      average concentration found of x, in jug/L.
C  =  True value for the concentration, in M9/L.
X  =  Average  recovery  found  for  measurements  of  samples  containing  a
      concentration of C, in M9/L.
a From 40 CFR Part 136 for Method 601.
b
  Estimates based upon the performance in  a single laboratory.
                               8010B - 13
    Revision 2
September 1994

-------
                       FIGURE  1.
GAS  CHROMATOGRAM OF HALOGENATED VOLATILE ORGANICS
                                 Column:
                                 Program:
                                 Detector:
1X SP-1000 on Carbopack-B
45°C-3 Minutes, 8'C/Hinute to 220°C
Hall 700-A Electrolytic Conductivity
                14
                                20
                                     n    74
                     HfTf MTHW TMf MHIUTUI
                       8010B -  14
                     Revision  2
                 September 1994

-------
                                      METHOD  8010B
             HALOGENATED VOLATILE  ORGANICS BY  GAS  CHROMATOGRAPHY
        Start
7.1  Introduce compounds
into gas chromatograph
  by direct injection or
    purge-and-trap
    (Method 5030)
      7.2 Set gas
    chromatograph
      condition.
     7.3 Calibrate
 (refer to Method 8000)
    7.4.1 Introduce
  volatile compounds
 into gas chromatograph
  by purge-and-trap or
    direct injection.
  7.4.2 Follow Method
   8000 for analysis
     sequence, etc.
7.4.4 Record volume
 purged or injected
  and peak sizes.
   7.4.5 Calculate
    concentration
(refer to Method 8000)
      7.4.6 Are
      analytical
    interferences
     suspected?
                                            7.4.7 Is
                                          response for
                                             a peak
                                           off-scale?
                                                                   7.4.6 Analyze using
                                                                    second GC column.
                               7.4.7 Dilute second
                                aliquot of sample.
                                       8010B  - 15
                                             Revision  2
                                        September  1994

-------
                                 METHOD 8011

              1,2-DIBROMOETHANE AND 1.2-DIBROMO-3-CHLOROPROPANE
                   BY MICROEXTRACTION AND GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   This method  is applicable to  the  determination of  the following
compounds in drinking water and ground water:
      Compound Name                                   CAS No.a


      1,2-Dibromoethane (EDB)                         106-93-4
      l,2-Dibromo-3-chloropropane (DBCP)               96-12-8


      8  Chemical  Abstract Services  Registry Number.

      1.2   For compounds and matrices other than those listed in Section 1.1,
the laboratory  must demonstrate  the  usefulness of  the method  by  collecting
precision  and  accuracy  data  on  actual  samples  and  provide  qualitative
confirmation of results by gas chromatography/mass spectrometry (GC/MS).

      1.3   The experimentally determined method  detection limits (MDL) for EDB
and DBCP were  calculated to be  0.01  /Lig/L.  The method has been shown  to  be
useful for  these analytes over a concentration range of approximately  0.03 to 200
jiig/L.  Actual detection limits are highly dependent upon the characteristics of
the gas chromatographic system, sample matrix,  and calibration.

      1.4   This method  is  restricted to  use  by or under  the Supervision  of
analysts experienced in the  use of gas chromatography and in the interpretation
of gas  chromatograms.  Each analyst must demonstrate the ability to generate
acceptable results with this method using the procedure described in Section 8.2.

      1.5   1,2-Dibromoethane   and    l,2-Dibromo-3-chloropropane    have   been
tentatively classified  as known  or  suspected  human  or  mammalian carcinogens.
Pure standard materials and stock standard solutions of these compounds should
be handled  in a hood.  A NIOSH/MESA approved toxic gas respirator should be worn
when the analyst handles high concentrations of these toxic compounds.


2.0   SUMMARY OF METHOD

      2.1   Thirty five ml of sample are extracted with 2 mL of hexane.  Two  fiL
of the  extract are  then injected  into a gas  chromatograph equipped  with  a
linearized electron capture detector for  separation and analysis. Aqueous matrix
spikes are  extracted and  analyzed  in an identical  manner as the  samples in order
to compensate for possible extraction  losses.

      2.2   The extraction  and analysis time  is  30  to 50  minutes  per sample

                                   8011  - 1                         Revision 0
                                                                     July 1992

-------
depending upon the analytical conditions chosen.  See Table 1 and Figure 1.

      2.3   Confirmatory evidence is obtained using a different column (Table 1).


3.0   INTERFERENCES

      3.1   Impurities  contained  in the  extracting solvent  (hexane)  usually
account for the majority of the analytical problems.  Reagent blanks should be
analyzed for each new bottle  of hexane before use.   Indirect daily checks on the
hexane are obtained by monitoring the reagent blanks.  Whenever an interference
is noted in the method or instrument blank, the  laboratory should reanalyze the
hexane.   Low  level  interferences  generally  can be  removed  by  distillation or
column chromatography, however,  it  is generally  more economical to obtain a new
source of hexane solvent.  Interference-free hexane is defined as containing less
than 0.01 /ug/L of the analytes.  Protect interference-free hexane by storing it
in an area known to be free of organochlorine solvents.

      3.2   Several  instances of  accidental sample  contamination have  been
attributed to  diffusion  of volatile organics through  the septum seal  into the
sample bottle during shipment and storage. Trip blanks must  be used to monitor
for this problem.

      3.3   This 1 iquid/1 iquid extraction technique extracts a wide boil ing range
of non-polar  organic compounds  and, in addition,  extracts  some polar organic
compounds.

      3.4   EDB at low concentrations may be  masked  by very high concentrations
of dibromochloromethane (DBCM), a common chlorinated drinking  water contaminant,
when using the confirmation column.


4.0   APPARATUS AND MATERIALS

      4.1   Microsyringe  - 10, 25,  and  100 juL  with a 2 in.  x 0.006 in. needle
(Hamilton 702N  or equivalent).

      4.2   Gas  Chromatograph

            4.2.1 The  GC must be capable  of  temperature programming and should
      be  equipped with a linearized electron capture detector and a capillary
      column  splitless injector.

            4.2.2 Columns

                  4.2.2.1     Column  A  - 0.32 mm  ID  x  30  m  fused silica
            capillary  with dimethyl silicone mixed phase  (Durawax-DX 3, 0.25 /xm
            film, or  equivalent).

                  4.2.2.2     Column B (confirmation column)  - 0.32 mm ID x  30  m
            fused silica capillary  with methyl polysiloxane phase (DB-1, 0.25 jum
            film, or  equivalent).

      4.3   Volumetric flasks, Class A  -  10  mL.

                                    8011  -  2                         Revision  0
                                                                     July  1992

-------
      4.4   Glass bottles - 15 mL, with Teflon lined screw caps or crimp tops.

      4.5   Analytical balance - 0.0001 g.

      4.6   Graduated cylinder - 50 mL.

      4.7   Transfer pi pet.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests. Unless otherwise
indicated, it is intended that all  reagents shall conform to the specifications
of the Committee on Analytical Reagents of the American  Chemical Society, where
such specifications are available. Other grades may  be used, provided it  is first
ascertained that the  reagent  is  of sufficiently  high  purity to permit its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Hexane, C6H14  - UV grade (Burdick and Jackson #216 or equivalent).

      5.4   Methyl alcohol, CH3OH - Demonstrated  to be free of analytes.

      5.5   Sodium chloride, NaCl - Pulverize  a batch  of NaCl  and place it in a
muffle  furnace  at room  temperature.    Increase  the temperature  to 400°C for
30 minutes.  Store in a capped bottle.

      5.6   1,2-Dibromoethane  (99%),   C2H4Br2,  (Aldrich  Chemical   Company,  or
equivalent).

      5.7   l,2-Dibromo-3-chloropropane  (99.4%),  C3H5Br2Cl,  (AMVAC   Chemical
Corporation, Los Angeles, California, or equivalent).

      5.8   Stock  standards  - These  solutions  may be  purchased  as certified
solutions or prepared from pure  standards using the following procedures:

            5.8.1 Place  about 9.8 mL  of methanol   into a 10 mL  ground glass
      stoppered volumetric flask.   Allow the  flask to stand, unstoppered, for
      about 10 minutes and weigh to the nearest 0.0001  g.

            5.8.2 Use  a  25 ptL  syringe  and  immediately add  two or more drops
      (» 10 juL)  of standard  to the flask.    Be  sure that  the  standard falls
      directly into the alcohol  without contacting the  neck of the flask.

            5.8.3 Reweigh, dilute to volume,  stopper,  and  then mix  by inverting
      the flask several  times.   Calculate the concentration in milligrams per
      liter (mg/L) from the net  gain in weight.  When compound purity is assayed
      to  be 96%  or greater,  the weight may be  used  without correction  to
      calculate the concentration of the stock standard.

            5.8.4 Store stock standards  in  15 mL bottles  equipped with Teflon
      lined screw-caps or crimp  tops.  Stock standards  are stable  for  at least

                                   8011 - 3                         Revision 0
                                                                     July 1992

-------
      four weeks when stored at 4°C and away from light.

      5.9   Intermediate  standard  -   Use   stock  standards   to  prepare  an
intermediate standard that contains both analytes in methanol.  The intermediate
standard should be  prepared  at a concentration that can  be easily diluted to
prepare aqueous calibration standards that will bracket the working concentration
range.   Store  the  intermediate  standard  with  minimal  headspace and  check
frequently for signs of deterioration or evaporation,  especially  just before
preparing calibration standards.  The storage time described for stock standards
also applies to the  intermediate standard.

      5.10  Quality control  (QC) reference sample - Prepare a QC reference sample
concentrate at 0.25 mg/L of both analytes from standards  from a different source
than the standards used for the stock standard.

      5.11  Check  standard  -  Add  an   appropriate  volume  of the intermediate
standard to an aliquot of organic-free reagent water in  a volumetric flask.  Do
not add more than  20 /zL of  an  alcoholic intermediate standard  to the water or
poor precision will  result.   Use a 25 /iL microsyringe  and  rapidly inject the
alcoholic  intermediate  standard into  the expanded  area of the  almost filled
volumetric flask.   Remove  the needle as quickly as possible after injection.  Mix
by inverting the flask several times.  Discard the contents contained in  the neck
of the flask.  Aqueous  calibration standards  should be prepared every 8 hours.


6.0   SAMPLE COLLECTION,  PRESERVATION, AND STORAGE

      6.1   See the  introductory material to  this  chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Recommended Chromatographic Conditions

      Two  gas  chromatography columns are  recommended.   Column A  is a highly
efficient column that provides separations for EDB and DBCP without interferences
from trihalomethanes.  Column A should be used as the primary analytical column
unless routinely  occurring  analytes  are  not adequately resolved.   Column B  is
recommended for use as a  confirmatory  column when GC/MS  confirmation is not
available.  Retention times  for EDB  and DBCP  on these columns are presented  in
Table 1.

      Column A:

      Injector  temperature:                     200°C.
      Detector  temperature:                     290°C.
      Carrier gas  (Helium)  Linear velocity:     25 cm/sec.
      Temperature  program:
            Initial  temperature:                40°C, hold for 4 min.
            Program:                            40°C to 190°C at 8°C/min.
            Final  temperature:                  190°C,  hold for  25  min.,   or
                                                until   all  expected  analytes
                                                have eluted.

                                   8011  -  4                         Revision 0
                                                                      July 1992

-------
See Figure 1 for a sample chromatogram and Table 1 for retention data.

Column B:

Injector temperature:                     200°C.
Detector temperature:                     290°C.
Carrier gas (Helium) Linear velocity:     25 cm/sec.
Temperature program:
      Initial temperature:                40°C,  hold for 4 min.
      Program:                            40°C to 270°C  at 10°C/nrin.
      Final temperature:                  270°C,   hold  for  10 min.,  or
                                          until   all  expected  analytes
                                          have eluted.

See Table 1 for retention data.

7.2   Calibration

      7.2.1 Prepare  at  least  five  calibration  standards.    One  should
contain EDB and DBCP at  a concentration near, but greater than, the method
detection limit  (Table  1) for each  compound.   The others  should  be at
concentrations  that bracket  the  range  expected in  the samples.   For
example,  if the  MDL is  0.01  M9/U  and a  sample  expected  to  contain
approximately 0.10 pg/L is to be analyzed, aqueous calibration standards
should be prepared at concentrations of 0.03 ;ug/L, 0.05 /xg/L, 0.10
0.15 M9/L,  and 0.20
      7.2.2 Analyze each calibration standard  and  tabulate peak height or
area  response versus  the  concentration  in  the  standard.    Prepare  a
calibration  curve  for each  compound.   Alternatively,  if the  ratio of
response  to  concentration (calibration factor)  is  a constant  over the
working  range (<  10% relative  standard  deviation),  linearity  can be
assumed and the average ratio or calibration factor can be used in place
of a calibration curve.

7.3   Sample preparation

      7.3.1 Remove samples and standards  from storage and allow them to
reach room temperature.

      7.3.2 For  samples  and  field  blanks contained  in 40  ml bottles,
remove the container  cap.   Discard a 5 ml volume using a 5 ml transfer
pipet.  Replace the container cap and weigh the container with contents to
the nearest  0.1  g and record this weight for  subsequent  sample volume
determination.

      7.3.3 For  calibration  standards,  check  standards,  QC  reference
samples,  and  blanks,  measure a  35  mL  volume  using  a 50  ml  graduated
cylinder  and transfer it to  a 40 mL sample container.

7.4   Extraction

      7.4.1 Remove the container cap and add 7 g  of NaCl to all samples.


                             8011 - 5                         Revision 0
                                                               July  1992

-------
      7.4.2 Recap the sample container and dissolve the NaCl by shaking by
hand for about 20 seconds.

      7.4.3 Remove the  cap and  using  a transfer  pipet,  add 2.0  ml of
hexane.  Recap and shake vigorously by hand for 1  minute.  Allow the water
and  hexane phases  to separate.    If  stored  at this  stage,  keep  the
container upside down.

      7.4.4 Remove the  cap and  carefully  transfer a  sufficient amount
(0.5-1.0 mL)  of  the hexane layer  into  a vial using  a disposable glass
pipet.

      7.4.5 Transfer  the  remaining hexane  phase,  being careful  not to
include any of the water phase, into a second vial.  Reserve this second
vial at 4°C for reanalysis if necessary.

7.5   Analysis

      7.5.1 Transfer  the  first sample  vial  to an  autosampler  set up to
inject  2.0  p.1   portions  into  the  gas  chromatograph  for  analysis.
Alternately,  2 /uL  portions  of samples,  blanks  and  standards  may be
manually injected,  using  the solvent flush technique,  although  an  auto
sampler is strongly recommended.

7.6   Determination of sample volume

      7.6.1 For samples and field blanks, remove the cap from the sample
container. Discard  the  remaining sample/hexane  mixture.   Shake  off the
remaining few drops using short, brisk wrist movements. Reweigh the empty
container  with original  cap  and calculate  the net weight of  sample by
difference to  the nearest 0.1 g.  This  net weight is equivalent to the
volume of water extracted.

7.7   Calculations

      7.7.1 Identify EDB and DBCP in the  sample chromatogram  by comparing
the retention time of the  suspect peak to retention times generated by the
calibration standards and the check standard.

      7.7.2 Use the  calibration curve  or calibration factor to  directly
calculate the uncorrected concentration  (Cj) of each analyte in the sample
(e.g. calibration factor  x response).

      7.7.3 Calculate the sample volume  (Vs)  as  equal  to the net  sample
weight:

      Vs (mL) = gross weight  (grams) - bottle tare (grams)

      7.7.4 Calculate the corrected sample  concentration as:

      Concentration  (jLtg/L)  =  C,  x 35
                                Vs

      7.7.5 Report the results for  the unknown  samples in fj.g/1.   Round the

                              8011  - 6                         Revision  0
                                                               July  1992

-------
      results to the nearest 0.01 /ig/L or two significant figures.


8.0   QUALITY CONTROL

      8.1   Each laboratory that uses this method is required to operate a formal
quality control program.

            8.1.1 The  laboratory  must make  an initial  determination of  the
      method detection limits and demonstrate the ability to generate acceptable
      accuracy and precision with this method.  This  is established as described
      in Section 8.2.

            8.1.2 In recognition of  laboratory advances that  are occurring in
      chromatography, the laboratory  is permitted certain options  to improve the
      separations  or  lower the  cost of measurements.    Each   time  such  a
      modification is made to the method,  the analyst is required to repeat the
      procedure in Section 7.1 and 8.2.

            8.1.3 The laboratory must analyze a reagent  and calibration blank to
      demonstrate  that  interferences  from  the  analytical  system are  under
      control every twenty  samples or per analytical batch,  whichever is more
      frequent.

            8.1.4 The laboratory must, on  an  ongoing basis, demonstrate through
      the  analyses  of QC  reference  samples  and  check  standards  that  the
      operation of the measurement system is in control.  The frequency of the
      check  standard  analyses  is  equivalent to  5% of all  samples  or  every
      analytical batch, whichever is more frequent.   On a weekly basis,  the QC
      reference sample must be run.

      8.2   To establish the ability to achieve low detection limits and generate
acceptable  accuracy and  precision,   the  analyst  must  perform  the  following
operations:

            8.2.1 Prepare seven samples each at a concentration of 0.03 jig/L.

            8.2.2 Analyze  the  samples according  to the  method  beginning  in
      Section 7.0.

            8.2.3 Calculate  the average  concentration (X)  in  ^g/L  and  the
      standard deviation  of the concentrations (s)  in  iig/L,  for each analyte
      using the seven results.  Then calculate the MDL  at 99% confidence level
      for seven replicates as 3.143s.

            8.2.4 For  each  analyte  in an  aqueous  matrix  sample,  X must  be
      between 60%  and  140% of the  true value.  Additionally,  the MDL may not
      exceed  the  0.03  jig/L spiked concentration.   If   both analytes  meet the
      acceptance criteria, the system performance  is acceptable and analysis of
      actual  samples can  begin.  If  either  analyte  fails  to meet a criterion,
      repeat  the  test.   It is recommended that the laboratory repeat the MDL
      determination on a regular basis.

      8.3   The laboratory must demonstrate on a frequency equivalent to 5% of

                                   8011 - 7                         Revision 0
                                                                     July 1992

-------
the sample load or once per analytical  batch, whichever is more frequent, that
the measurement  system is in  control  by analyzing  a  check standard  of both
analytes at 0.25 jig/L.

            8.3.1 Prepare  a  check  standard  (0.25  [ig/L)  by  diluting  the
      intermediate standard with water to 0.25 jig/L.

            8.3.2 Analyze the sample according to Section 7.0 and calculate the
      recovery for each analyte.  The recovery must be between 60% and 140% of
      the expected value  for  aqueous matrices.   For non-aqueous matrices, the
      U.S. EPA will set criteria after more interlaboratory data are gathered.

            8.3.3 If  the  recovery  for  either  analyte  falls  outside  the
      designated range,  the  analyte fails the acceptance  criteria.   A second
      calibration verification standard containing each analyte that failed must
      be analyzed.  Repeated  failure, however, will  confirm a general problem
      with the  measurement system.   If this  occurs,  locate  and correct the
      source of the problem and repeat the test.

      8.4   On a weekly  basis,  the  laboratory must  demonstrate the  ability to
analyze a QC reference sample.

            8.4.1 Prepare a QC reference sample at 0.10 ng/L by diluting the QC
      reference sample concentrate  (Section 5.9).

            8.4.2 For each analyte  in an  aqueous matrix,  the recovery must be
      between 60% and  140% of  the expected value.  When either analyte fails the
      test, the analyst must repeat the  test  only for that  analyte which failed
      to meet the criteria.  Repeated failure, however, will confirm a general
      problem with the measurement  system or faulty samples and/or standards.
      If this occurs,  locate  and  correct  the source of the problem and repeat
      the test.  For non-aqueous matrices,  the U.S.  EPA will set criteria after
      more interlaboratory data are gathered.

      8.5   Instrument  performance  -  Check the  performance  of the entire
analytical system daily using data gathered from analyses of blanks, standards,
and replicate samples.

            8.5.1 Peak tailing significantly in excess  of that  shown in the
      chromatogram (Figure 1)  must be corrected.  Tailing  problems  are generally
      traceable to active  sites on  the GC column or to the detector operation.

            8.5.2 Check  the precision between replicate  analyses.  A  properly
      operating  system  should  perform  with an   average relative   standard
      deviation  of  less  than  10%.   Poor precision  is generally  traceable to
      pneumatic leaks, especially at the injection port.


9.0   METHOD PERFORMANCE

      9.1   Method detection limits are  presented in Table 1.  Single laboratory
accuracy and precision at  several concentrations in tap water  are presented in
Table 2.


                                    8011 - 8                         Revision  0
                                                                     July  1992

-------
      9.2   In a preservation study extending  over a 4 week period, the average
percent recoveries and relative standard deviations presented  in Table 3 were
observed for organic-free reagent water (acidified),  tap water and ground water.
The results  for acidified  and non-acidified  samples  were not  significantly
different.
10.0  REFERENCES

1.    Optimization of Liquid-Liquid Extraction Methods for Analysis of Organics
      in Water, EPA-600/S4-83-052, 1984.

2.    Henderson, J.E.; Peyton,  G.R.;  Glaze, W.H.  Identification  and Analysis of
      Organic Pollutants in Water; Keith, L.H.,  Ed; Ann Arbor Sci.: Ann Arbor,
      MI; 1976.

3.    Richard J.J.; Junk, G.A. Journal AWWA 1977, 69, 62.

4.    Budde, W.L.; Eichelberger, J.W. Organic Analyses Using Gas Chromatography-
      Mass Spectrometry; Ann Arbor Science: Ann Arbor,  MI; 1978.

5.    Glaser, J.A.; et al.  Environmental  Science  and Technology  1981, 15, 1426.

6.    Methods for the  Determination  of  Organic Compounds in  Finished Drinking
      Water and Raw Source Water; U.S. Environmental  Protection Agency. Office
      of  Research  and  Development.  Environmental   Monitoring  and  Support
      Laboratory.  ORD Publication Offices  of Center for  Environmental Research
      Information:  Cincinnati, OH 1986.
                                   8011 - 9                         Revision 0
                                                                     July 1992

-------
                                   TABLE 1.
                CHROMATOGRAPHIC  CONDITIONS  AND  METHOD DETECTION
                 LIMITS  (MDL)  FOR  1,2-DIBROMOETHANE  (EDB)  AND
                      1.2-DIBROMO-3-CHLOROPROPANE (DBCP)
Analyte
Retention Time, Minutes

Column A    Column B      MDL (jig/L)
EDB

DBCP
 9.5

17.3
 8.9

15.0
0.01

0.01
Column A:  Durawax-DX 3

Column B:  DB-1
                                   TABLE 2.
                   SINGLE LABORATORY ACCURACY AND PRECISION
                         FOR EDB AND DBCP IN TAP WATER



Analyte
EDB


DBCP



Number
of
Samples
7
7
7
7
7
7

Spike
Concentration
(M9A)
0.03
0.24
50.0
0.03
0.24
50.0

Average
Recovery
(%)
114
98
95
90
102
94
Relative
Standard
Deviation
(%)
9.5
11.8
4.7
11.4
8.3
4.8
                                   8011 - 10
                                Revision 0
                                 July  1992

-------
                             TABLE 3.
                ACCURACY AND PRECISION AT 2.0
                    OVER A 4-WEEK STUDY PERIOD


Analyte
EDB




DBCP






Matrix1
RW-A
GW
GW-A
TW
TW-A
RW-A
GW
GW-A
TW
TW-A

Number
of Samples
16
15
16
16
16
16
16
16
16
16
Average
Accuracy
(% Recovery)
104
101
96
93
93
105
105
101
95
94
Relative
Std. Dev.
(%)
4.7
2.5
4.7
6.3
6.1
8.2
6.2
8.4
10.1
6.9
RW-A  =     Organic-free reagent water at pH 2
GW    =     Ground water, ambient pH
GW-A  =     Ground water at pH 2
TW    =     Tap water, ambient pH
TW-A  =     Tap water at pH 2
                             8011  -  11                         Revision 0
                                                               July 1992

-------
                             FIGURE 1.
          SAMPLE  CHROMATOGRAM  FOR  EXTRACT OF WATER SPIKED
                  AT 0.114 M9/L WITH EDB AND DBCP
                                       COLUMN:  Fused silica capillary
                                       LIQUID PHASE:  Durawax-DX3
                                       FILM THICKNESS:  0.25 urn
                                       COLUMN DIMENSIONS:  30 M x 0.317
           10
     I	I	I     111     I    I    I
                                                     i     i
2    4    •   •    10  11  14   It   It  20  22   24   2t
                          TIME (MIN)
                              8011 - 12
Revision 0
 July 1992

-------
                      METHOD 8011
1,2-DIBROMOETHANE  AND  l,2-DIBROMO-3-CHLOROPROPANE
     BY MICROEXTRACTION AND GAS  CHROMATOGRAPHY
       7 2 Check
      ins t rutnent
     performance
                          7 4  1 Add
                           NaCl to
                          samples
 743 Add
hexane and
  ex t ract
  sample
                          744 Put
                           part of
                         extract in
                            vial
                         745 Save
                         remainder of
                         extract for
                          possible
                         reanalysis
                  7 5 Analyze
                    by GC
7  6 Determine
  sample
  volume
                  7 7 Calculate
                 concent rations
                     Stop
                        8011  -  13
                                     Revision  0
                                      July  1992

-------
                                 METHOD 8015A

            NONHALOGENATED VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      I.I   Method  8015  is  used to  determine the  concentration of  various
nonhalogenated  volatile  organic compounds.   The  following compounds  can  be
determined by this method:
                                                     Appropriate Technique
                                                                      Direct
Compound Name                        CAS No.a       Purge-and-Trap     Injection
Di ethyl ether
Ethanol
Methyl ethyl ketone (MEK)
Methyl isobutyl ketone (MIBK)
60-29-7
64-17-5
78-93-3
108-10-1
b
i
PP
PP
b
b
b
b
     a  Chemical Abstract Services Registry Number.
     b  Adequate response using this technique
     i  Inappropriate technique for this analyte
     pp Poor purging efficiency, resulting in high EQLs


2.0   SUMMARY OF METHOD

      2.1   Method 8015 provides gas chromatographic conditions for the detection
of certain nonhalogenated volatile organic compounds.  Samples may be introduced
into the GC using direct injection or purge-and-trap (Method 5030).  Ground water
samples must be analyzed by Method 5030.  A temperature program is used in the
gas chromatograph to separate the organic compounds.  Detection is achieved by
a flame ionization detector (FID).

      2.2   The method provides an  optional gas chromatographic column that may
be helpful in resolving the analytes from co-eluting non-target compounds and for
analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Method 5030 and 8000.

      3.2   Samples  can   be  contaminated by  diffusion  of  volatile  organics
(particularly chlorofluorocarbons  and  methylene chloride)  through  the  sample
container  septum  during  shipment  and   storage.   A  trip blank  prepared  from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.


                                   8015A - 1                         Revision 1
                                                                     July 1992

-------
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  Chromatograph   -  Analytical  system complete  with  gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and  all  required  accessories, including  detectors,  column
      supplies, recorder, gases,  and syringes.  A data system  for measuring peak
      heights and/or peak areas is recommended.

            4.1.2 Columns

                  4.1.2.1     Column  1 - 8 ft x 0.1  in. ID  stainless steel or
            glass column  packed  with  1% SP-1000 on  Carbopack-B  60/80  mesh or
            equivalent.

                  4.1.2.2     Column  2 - 6 ft  x  0.1  in.  ID stainless steel or
            glass  column  packed  with  n-octane on   Porasil-C  100/120  mesh
            (Durapak) or equivalent.

            4.1.3 Detector - Flame ionization (FID).

      4.2   Sample  introduction   apparatus  -  Refer  to  Method  5030 for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes - A 5 ml Luerlok glass hypodermic  and a 5  ml, gas-tight with
shutoff valve.

      4.4   Volumetric  flasks, Class  A  -  Appropriate  sizes  with  ground glass
stoppers.

      4.5   Microsyringes -  10 and 25 /zL with  a 0.006 in. ID needle (Hamilton
702N or equivalent) and a 100 /LtL.

      4.6   Analytical  balance - 0.0001 g.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in  all tests. Unless otherwise
indicated, it is intended that all  reagents shall conform  to  the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are available. Other grades  may be  used, provided it is first
ascertained that the  reagent  is  of sufficiently high purity  to permit its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water,  as defined  in Chapter One.

      5.3   Methanol, CH3OH.   Pesticide  quality or equivalent.  Store away from
other solvents.

      5.4   Stock standards - Stock solutions may be  prepared  from pure standard
materials  or  purchased  as  certified  solutions.   Prepare stock  standards in

                                   8015A -  2                         Revision 1
                                                                     July 1992

-------
methanol using assayed liquids.

            5.4.1 Place about 9.8 ml of methanol  in a 10 ml tared, ground-glass
      stoppered volumetric flask.   Allow  the  flask to  stand,  unstoppered, for
      about 10 minutes or until all alcohol wetted surfaces have dried.  Weigh
      the flask to the nearest 0.0001 g.

            5.4.2 Using a 100 /iL syringe,  immediately add two or more drops of
      assayed reference material to  the  flask; then  reweigh.   The liquid must
      fall directly into the alcohol without contacting the neck of the flask.

            5.4.3 Reweigh, dilute to volume, stopper, and then mix by inverting
      the flask several times.   Calculate the concentration in milligrams per
      liter (mg/L) from the net gain  in weight. When  compound purity is assayed
      to  be  96%  or  greater,  the  weight  may be used  without  correction  to
      calculate the concentration of the stock standard.  Commercially prepared
      stock standards may be  used at any concentration if they are certified by
      the manufacturer or by an independent source.

            5.4.4 Transfer the  stock standard solution  into  a  bottle  with  a
      Teflon lined screw-cap.  Store, with minimal headspace,  at -10°C  to  -20°C
      and protect from light.

            5.4.5 Standards  must  be  replaced after  6 months,  or  sooner  if
      comparison with check standards indicates a problem.

      5.5   Secondary dilution standards - Using  stock standard solutions, pre-
pare  in  methanol  secondary  dilution standards,   as  needed, that  contain the
compounds of interest, either singly or mixed  together.  The secondary dilution
standards should be prepared at concentrations such that the  aqueous calibration
standards  prepared   in  Section  5.5 will  bracket the  working  range  of the
analytical system.  Secondary dilution standards should be stored with minimal
headspace for volatiles and should be checked frequently for signs of degradation
or evaporation, especially just  prior  to preparing  calibration standards from
them.

      5.6   Calibration standards  - Calibration standards at a minimum of five
concentrations are prepared  in water from the secondary  dilution of the  stock
standards.  One  of  the concentrations should be  at  a  concentration near, but
above,  the  method   detection  limit.    The  remaining  concentrations  should
correspond to  the expected range  of concentrations  found  in  real  samples or
should define the working range  of the GC.  Each standard should contain each
analyte for detection by this method (e.g. some or all  of the compounds listed
in Section 1.1 may be included).   In order to  prepare accurate aqueous standard
solutions, the following precautions must  be observed:

            5.6.1 Do  not  inject  more than  20 jiiL of alcoholic  standards into
      100 ml of water.

            5.6.2 Use  a  25  /xL   Hamilton  702N  microsyringe  or  equivalent
      (variations  in needle  geometry will  adversely  affect  the  ability  to
      deliver reproducible volumes of methanolic standards  into water).

            5.6.3 Rapidly  inject  the  alcoholic  standard  into  the  filled

                                   8015A - 3                         Revision 1
                                                                     July 1992

-------
      volumetric flask.  Remove the  needle  as fast as possible after injection.

            5.6.4 Mix aqueous standards by  inverting the flask three times only.

            5.6.5 Fill the sample syringe from the standard solution contained
      in the expanded  area of  the flask  (do  not  use  any solution contained in
      the neck of the flask).

            5.6.6 Never  use  pipets  to dilute  or transfer samples  or aqueous
      standards.

            5.6.7 Aqueous standards  are not stable and should  be discarded after
      1 hour, unless properly sealed and stored.   The aqueous standards can be
      stored up to 24 hours,  if held in sealed vials with zero headspace.

      5.7   Internal standards (if internal standard calibration is used) - To
use this approach, the analyst must select one or more internal  standards that
are similar in  analytical behavior  to the  compounds  of interest.  The analyst
must further demonstrate that  the measurement  of the internal  standard is not
affected by method  or  matrix interferences.   Because of these limitations, no
internal standard can be suggested that is applicable to all  samples.

            5.7.1 Prepare  calibration   standards   at   a  minimum   of   five
      concentrations for each parameter of  interest as described  in Section 5.6.

            5.7.2 Prepare a  spiking solution containing each of the internal
      standards using the procedures described in Sections 5.4 and 5.5.  It is
      recommended  that  the   secondary  dilution  standard  be  prepared  at  a
      concentration  of  15  ng/juL of each  internal   standard  compound.    The
      addition  of  10 fj,L of  this  standard  to 5.0 ml  of  sample  or calibration
      standard would be equivalent to 30
            5.7.3 Analyze each  calibration  standard  according to Section 7.0,
      adding  10 /j,l  of  internal  standard  spiking  solution  directly  to  the
      syringe.

      5.8   Surrogate standards - The analyst should monitor both the performance
of the analytical  system and the effectiveness of the method in dealing with each
sample matrix by spiking each sample,  standard,  and water  blank with  one or two
surrogate compounds  recommended to  encompass the  range of temperature program
used in this method.   From stock standard solutions prepared  as in Section 5.4,
add a volume to give 750 jug of  each surrogate to 45 ml of water contained in a
50  ml  volumetric  flask,  mix,  and  dilute  to  volume  for a  concentration  of
15 ng//iL.  Add  10 /xL of this  surrogate  spiking  solution directly into the 5 ml
syringe with  every sample and  reference standard analyzed.   If the internal
standard calibration procedure is used, the  surrogate  compounds  may be added
directly to the internal standard spiking solution (Section 5.7.2).


6.0   SAMPLE COLLECTION, PRESERVATION,  AND  HANDLING

      6.1   See the  introductory  material  to this  Chapter,  Organic Analytes,
Section 4.1.


                                   8015A - 4                         Revision 1
                                                                      July 1992

-------
7.0   PROCEDURE

      7.1   Volatile compounds are introduced  into the gas chromatograph either
by direct injection or purge-and-trap  (Method  5030).   Method 5030 may be used
directly on ground  water samples or low-concentration  contaminated soils and
sediments.   For high-concentration soils or sediments, methanolic extraction, as
described in Method  5030,  may be necessary prior to  purge-and-trap analysis.
Method 5030 also provides guidance on the  analysis of aqueous miscible and non-
aqueous miscible liquid wastes (see Section 7.4.1.1).

      7.2   Chromatographic conditions (Recommended)

            7.2.1 Column 1

            Carrier gas (Helium) flow rate:     40 mL/min
            Temperature program:
                  Initial temperature:           45°C,  hold for 3  minutes
                  Program:                      45°C to 220°C at 8°C/min
                  Final temperature:            220°C,  hold for 15 minutes.

            7.2.2 Column 2

            Carrier gas (Helium) flow rate:     40 mL/min
            Temperature program:
                  Initial temperature:           50°C,  hold for 3  minutes
                  Program:                      50°C to 170°C at 6°C/min
                  Final temperature:            170°C,  hold for 4  minutes.

      7.3   Calibration - Refer to Method 8000 for proper calibration  techniques.

            7.3.1 Calibration must take place using the same sample introduction
      method that will be used to analyze actual samples  (see Section 7.4.1).

            7.3.2 The  procedure  for internal  or  external  calibration  may be
      used.  Refer to  Method 8000 for a description of  each of these  procedures.

      7.4   Gas Chromatographic analysis

            7.4.1 Introduce volatile compounds into  the  gas  chromatograph using
      either Method 5030  (purge-and-trap method) or the direct injection method.
      If the  internal standard  calibration  technique  is  used,  add  10 jiiL of
      internal standard to the sample prior to purging.

                  7.4.1.1     Direct injection  -  In very limited applications
            (e.g. aqueous process wastes), direct injection of the  sample into
            the GC  system with  a 10 juL syringe may be  appropriate.  One such
            application is for verification of the alcohol content  of an aqueous
            sample prior to determining if the sample is ignitable  (Methods 1010
            or 1020).  In this  case,  it  is suggested  that direct injection be
            used. The detection  limit is very high (approximately 10,000 M9/L);
            therefore, it  is  only permitted when concentrations  in excess of
            10,000 jug/L are expected or for water-soluble compounds  that do not
            purge.  The system must be calibrated by  direct injection (bypassing
            the purge-and-trap device).

                                   8015A - 5                         Revision  1
                                                                     July 1992

-------
                  Non-aqueous miscible wastes may  also be analyzed  by  direct
            injection if the  concentration of target  analytes  in the  sample
            falls within the calibration range.   If dilution of the  sample  is
            necessary,  follow the  guidance for  High Concentration samples  in
            Method 5030, Section 7.3.3.2.

            7.4.2 Method 8000 provides instructions on the  analysis  sequence,
      appropriate dilutions,  establishing  daily retention  time windows,  and
      identification criteria.   Include a mid-concentration standard after each
      group of 10 samples in the analysis  sequence.*

            7.4.3 Record the sample volume  purged or injected and the resulting
      peak sizes (in area units or peak heights).

            7.4.4 Calculation of concentration is  covered in Method 8000.

            7.4.5 If analytical  interferences  are suspected, or for the purpose
      of confirmation,  analysis using the  second GC column is recommended.

            7.4.6 If the response for a peak is off-scale,  prepare a dilution of
      the sample with water.  The dilution  must be performed on a second aliquot
      of the sample which has been properly sealed and stored prior to use.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One  for specific quality control  procedures and
Method 8000 for gas chromatographic procedures.   Quality control to ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2   Quality control  required to validate the GC system operation is found
in Method 8000, Section 8.6.

      8.3   Calculate surrogate standard recovery  on  all  samples,  blanks, and
spikes.  Determine if recovery is within limits (limits  established by performing
QC procedure outlined in Method 8000, Section 8.10).

            8.3.1 If recovery is not within limits, the following is required:

            •     Check  to  be  sure that there are  no  errors in calculations,
                  surrogate  solutions,  and internal  standards.    Also,  check
                  instrument performance.

            •     Recalculate the  data and/or reanalyze  the  extract  if any of
                  the above checks reveal  a problem.

            •     Re-extract and re-analyze the  sample if none of the above are
                  a problem or flag the data as "estimated concentration".


9.0   METHOD PERFORMANCE

      9.1   The accuracy and precision obtained will be determined  by  the sample
matrix,  sample  introduction technique, and calibration procedures used.

                                   8015A -  6                        Revision  1
                                                                     July 1992

-------
      9.2   Specific method  performance  information  will  be  provided  as  it
becomes available.
10.0  REFERENCES

1.    Bellar, T.A.,  and J.J.  Lichtenberg, Determining  Volatile Organics  at
                              '   '   "                    J. Amer. Water Works
Bellar,  T.A.,  and  J.J.  Lichtenberg,  Determining
Microgram-per-Liter Levels  by  Gas Chromatography,
Assoc., 66(12). pp. 739-744 (1974).
2.    Bellar, T.A., and J.J. Lichtenberg, Semi-Automated Headspace Analysis of
      Drinking Waters  and  Industrial  Waters for  Purgeable Volatile  Organic
      Compounds,  in Van Hall, ed., Measurement of  Organic  Pollutants in Water
      and Wastewater, ASTM STP 686, pp. 108-129,  1979.

3.    Development and Application of Test Procedures for Specific Organic Toxic
      Substances in Wastewaters:   Category 11 -  Purgeables  and  Category 12 -
      Acrolein,  Acrylonitrile,   and  Dichlorodifluoromethane,  Report for  EPA
      Contract 68-03-2635 (in preparation).
                                  8015A  - 7                         Revision 1
                                                                     July 1992

-------
                          METHOD 8015A
NONHALOGENATED  VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
           Start
          7 2 Set
       chroma tographic
        condi11ona
        7 3 Calibrate
         (refer  to
        Method 8000)
       741 Introduce
       sample into CC
         by direct
       infection or
       purge-and-trap
       742 Follow
        Method  8000
       for analysis
         sequence,
           etc
 744 Record
 volume purged
     or
 injected,and
  peak sizes
7 4  5 Calculate
concentrations
   (refer to
 Method 8000)
   7 4 6 Are
  analytical
 interferences
  suspected?
 7  4 7 Is peak
 response off
   scale?
746 Analyze
sample using
  second CC
   column
747 Dilute
   second
 aliquot of
   sample
                             8015A  -  8
                                           Revision  1
                                            July  1992

-------
                                 METHOD 8020A

               AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8020  is  used to  determine the  concentration of  various
aromatic volatile organic compounds.   The following compounds can be determined
by this method:
                                                   Appropriate Technique
                                                                    Direct
Compound Name                        CAS  No.a       Purge-and-Trap  Injection
Benzene
Chlorobenzene
1,2-Dichlorobenzene
1, 3 -Di chlorobenzene
1,4-Dichlorobenzene
Ethyl benzene
Toluene
Xylenes
a Chemical Abstract
b adequate response
71-43-2
108-90-7
95-50-1
541-73-1
106-46-7
100-41-4
108-88-3

Services Registry Number.
by this technique.
b
b
b
b
b
b
b
b


b
b
b
b
b
b
b
b


      1.2   Table 1  lists the method detection limit for each target analyte in
organic-free reagent water.  Table 2 lists the estimated quantitation limit (EQL)
for other matrices.
2.0   SUMMARY OF METHOD

      2.1   Method 8020 provides chromatographic conditions for the detection of
aromatic volatile compounds.  Samples can be introduced into the GC using direct
injection  or purge-and-trap  (Method 5030).    Ground water  samples  must  be
determined  using  Method  5030.    A  temperature  program  is  used  in  the  gas
chromatograph to  separate  the  organic  compounds.   Detection is  achieved  by a
photo-ionization detector (PID).

      2.2   If interferences are encountered,  the method provides an optional gas
chromatographic column that may be helpful in  resolving  the  analytes  from the
interferences and for analyte confirmation.


3.0   INTERFERENCES

      3.1   Refer to Method 5030  and  8000.
                                  8020A  -  1                         Revision 1
                                                                September 1994

-------
      3.2   Samples  can  be  contaminated  by  diffusion  of volatile  organics
(particularly chlorofluorocarbons  and  methylene chloride) through  the sample
container septum during shipment and storage.  A field sample blank prepared from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1  Gas Chromatograph  -  Analytical  system complete with  gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and all  required accessories, including detectors,  column
      supplies,  recorder, gases, and syringes.  A data  system for measuring peak
      heights and/or peak areas is recommended.

            4.1.2  Columns

                   4.1.2.1     Column 1:  6 ft x 0.082 in  ID #304 stainless steel
            or glass  column packed with  5% SP-1200  and 1.75% Bentone-34  on
            100/120 mesh  Supelcoport,  or equivalent.

                   4.1.2.2     Column 2:  8 ft  x 0.1 in  ID stainless  steel  or
            glass  column  packed with  5%  l,2,3-Tris(2-cyanoethoxy)propane  on
            60/80 mesh Chromosorb  W-AW, or equivalent.

            4.1.3  Detector - Photoionization (PID)  (h-Nu  Systems,  Inc.  Model
      PI-51-02 or equivalent).

      4.2   Sample  introduction  apparatus  -  Refer to  Method 5030  for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes - A 5 mL Luerlok glass hypodermic and a 5 mL, gas-tight with
shutoff valve.

      4.4   Volumetric flask,  Class  A  -  Appropriate  sizes with ground  glass
stoppers.

      4.5   Microsyringe  -  10 and 25 p.1 with a 0.006  in ID needle (Hamilton 702N
or equivalent) and a 100  juL.

      4.6   Analytical balance - 0.0001 g.


5.0   REAGENTS

      5.1   Organic-free  reagent water.   All references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.2   Methanol  (CH3OH)  - pesticide quality or equivalent.  Store away from
other solvents.


                                   8020A -  2                         Revision 1
                                                                September 1994

-------
      5.3   Stock standards -  Stock solutions may be prepared  from pure standard
materials  or  purchased  as  certified solutions.   Prepare stock  standards in
methanol  using  assayed  liquids.    Because  of  the  toxicity  of  benzene  and
1,4-dichlorobenzene, primary dilutions of these materials  should be prepared in
a hood.

            5.3.1  Place about 9.8 mL of methanol  in a  10 ml tared ground glass
      stoppered  volumetric flask.  Allow the  flask to  stand, unstoppered,  for
      about 10 min or until all alcohol wetted surfaces have dried.  Weigh the
      flask to the nearest 0.0001 g.

            5.3.2  Using a 100 /zL syringe,  immediately  add  two or more drops of
      assayed reference  material  to  the  flask;  then  reweigh.   The liquid must
      fall directly into the alcohol  without contacting the neck of the flask.

            5.3.3  Reweigh, dilute to volume, stopper, and then mix by inverting
      the  flask  several  times.   Calculate  the  concentration  in  milligrams  per
      liter (mg/L) from the net gain  in weight.   When  compound purity is assayed
      to  be  96%  or  greater,  the weight  may  be  used without  correction to
      calculate the concentration of  the stock  standard.  Commercially prepared
      stock standards may be used at  any concentration  if they are certified by
      the manufacturer or by an independent source.

            5.3.4  Transfer the  stock standard solution  into  a  Teflon-sealed
      screw-cap bottle.  Store,  with  minimal  headspace, at  4°C and protect from
      light.

            5.3.5  All  standards must be replaced after 6 months, or sooner if
      comparison with check standards indicates a problem.

      5.4   Secondary  dilution  standards:   Using  stock  standard  solutions,
prepare in methanol secondary dilution  standards,  as  needed,  that contain  the
compounds of interest,  either singly  or  mixed together.  The secondary dilution
standards should be prepared at concentrations such that the aqueous calibration
standards  prepared in  Section  5.5  will  bracket the working  range  of  the
analytical system.  Secondary dilution standards should be stored with minimal
headspace for  volatiles and should be  checked frequently for signs of degradation
or evaporation,  especially just  prior to preparing calibration  standards from
them.

      5.5   Calibration standards:  Calibration standards at  a minimum of five
concentrations are prepared  in  organic-free reagent water from  the secondary
dilution  of the  stock standards.   One of  the  concentrations  should be at  a
concentration  near,  but  above,  the  method  detection  limit.    The  remaining
concentrations should correspond to the expected range of concentrations found
in real samples  or  should define the working range of the GC.   Each  standard
should contain each analyte for detection by this method (e.g.,  some or all of
the compounds listed in the target analyte list may be included).  In order to
prepare accurate aqueous standard solutions, the following precautions must be
observed.
                                   8020A  -  3                         Revision 1
                                                                September 1994

-------
            5.5.1  Do not inject more  than  20 /j.1 of  alcoholic  standards into
      100 ml of organic-free reagent water.

            5.5.2  Use  a 25   /iL   Hamilton  702N  microsyringe  or  equivalent
      (variations  in needle  geometry  will  adversely affect  the  ability  to
      deliver reproducible volumes of methanolic standards into water).

            5.5.3  Rapidly  inject   the  alcoholic  standard  into  the  filled
      volumetric flask.   Remove the needle as  fast as possible after injection.

            5.5.4  Mix aqueous  standards by inverting the flask three times only.

            5.5.5  Fill  the  sample  syringe from the standard solution contained
      in the expanded area of  the  flask  (do not  use  any  solution contained in
      the neck of the flask).

            5.5.6  Never use pipets  to dilute or transfer  samples or aqueous
      standards.

            5.5.7  Aqueous standards  are not stable and should be discarded after
      1 hr, unless  properly  sealed and stored.   The  aqueous  standards can be
      stored up to 24 hr, if held in sealed vials with zero headspace.

      5.6   Internal standards (if internal  standard  calibration is used):  To
use this approach, the analyst must select one or more internal  standards that
are similar in  analytical behavior  to  the compounds  of  interest.   The analyst
must further demonstrate that  the  measurement of  the  internal  standard is not
affected by method  or matrix  interferences.   Because  of  these limitations, no
internal  standard  can   be   suggested   that   is  applicable  to  all  samples.
Alpha,alpha,alpha-trifluorotoluene has  been  used successfully  as  an internal
standard.

            5.6.1  Prepare  calibration   standards   at   a  minimum  of  five
      concentrations for each  parameter of interest as described  in  Section 5.5.

            5.6.2  Prepare a spiking solution containing  each  of the internal
      standards using the procedures described in Sections 5.3 and 5.4.  It is
      recommended  that   the   secondary dilution  standard  be  prepared  at  a
      concentration of 15 mg/L  of each internal standard compound.   The addition
      of 10 IJ.L of this  standard to  5.0 ml of sample or  calibration standard
      would be equivalent to 30 M9/L.

            5.6.3  Analyze each calibration standard  according to  Section 7.0,
      adding  10 /uL  of  internal   standard  spiking  solution  directly  to  the
      syringe.

      5.7   Surrogate standards:  The analyst should monitor both  the performance
of the analytical  system  and the effectiveness  of the method in dealing  with each
sample matrix by spiking each  sample,  standard,  and organic-free reagent water
blank with surrogate compounds (bromochlorobenzene,  bromofluorobenzene, 1,1,1-
trifluorotoluene,  fluorobenzene,  and  difluorobenzene are  recommended) which
encompass the range of the temperature  program used in this method.  From stock


                                   8020A -  4                         Revision 1
                                                                September 1994

-------
standard solutions prepared  as  in  Section  5.3,  add  a volume to give 750 /ug of
each surrogate  to 45 ml  of organic-free  reagent water contained  in  a 50 ml
volumetric flask, mix,  and  dilute  to volume for a  concentration  of 15 ng/juL.
Add 10  jttL of this surrogate  spiking solution directly into the 5 ml  syringe with
every  sample  and reference  standard  analyzed.     If  the  internal   standard
calibration procedure is used,  the  surrogate compounds  may  be added  directly to
the internal standard spiking solution  (Section 5.6.2).


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See  the  introductory material  to this  chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into  the gas chromatograph either
by direct injection or  purge-and-trap  (Method 5030).   Method 5030 may be used
directly on  ground  water samples or low-concentration  contaminated soils and
sediments.   For medium-concentration soils  or sediments, methanolic extraction,
as described in Method 5030,  may be  necessary prior to purge-and-trap analysis.
Method 5030 also provides guidance on the analysis  of aqueous miscible and non-
aqueous miscible liquid wastes  (see Section  7.4.1.1 below).

      7.2   Gas chromatography conditions (Recommended):

            7.2.1  Column 1:

            Carrier gas (He)  flow rate:  36 mL/min
            For lower boiling compounds:
                   Initial temperature:  50°C, hold for 2 min;
                   Temperature program:  50°C  to 90°C  at  6°C/min,  hold until
                                         all compounds  have eluted.
            For higher boiling range of compounds:
                   Initial temperature:  50°C, hold for 2 min;
                   Temperature program:  50°C  to 110°C  at  3°C/min,  hold until
                                         all compounds  have eluted.

            Column 1  provides outstanding   separations  for  a wide  variety  of
      aromatic hydrocarbons.   Column 1  should be  used as the primary analytical
      column because of its unique ability to resolve para-, meta-, and ortho-
      aromatic isomers.

            7.2.2  Column 2:

            Carrier gas (He)  flow rate:  30 mL/min
            Initial  temperature:         40°C, hold for 2 min;
            Temperature program:         40°C  to 100°C  at  2°C/min,  hold until
                                         all compounds  have eluted.
                                   8020A  -  5                         Revision 1
                                                                September 1994

-------
            Column 2, an  extremely  high polarity column, has been  used  for a
      number of years to resolve aromatic hydrocarbons from alkanes in complex
      samples.  However, because resolution between some of the aromatics is not
      as efficient as with Column 1, Column 2 should be used as a confirmatory
      column.

      7.3   Calibration:  Refer to Method 8000 for proper calibration techniques.
Use Table 1  and  especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

            7.3.1  Calibration must take place using the same sample introduction
      method that will be used to analyze actual samples (see Section 7.4.1).

            7.3.2  The procedure  for  internal  or external calibration may be
      used.   Refer to Method 8000 for a description of  each of these procedures.

      7.4   Gas chromatographic  analysis:

            7.4.1  Introduce volatile compounds into the gas chromatograph using
      either Method 5030 (purge-and-trap method) or the direct injection method.
      If the  internal standard  calibration  technique is used,  add 10 /jL of
      internal standard to the sample prior to purging.

                   7.4.1.1    Direct injection:   In very limited applications
            (e.g., aqueous process wastes), direct  injection of the sample into
            the  GC  system  with   a  10  /nL  syringe  may  be  appropriate.    The
            detection limit  is very high (approximately 10,000 M9/L); therefore,
            it is  only permitted when  concentrations  in  excess of  10,000  /zg/L
            are expected  or  for water  soluble compounds that do not purge.   The
            system must  be calibrated  by direct  injection (bypassing the purge-
            and-trap device).

                   Non-aqueous miscible wastes may  also  be  analyzed by direct
            injection if  the concentration  of  target analytes  in  the sample
            falls within the calibration range.   If dilution  of  the sample is
            necessary, follow the guidance for  High  Concentration  samples in
            Method 5030,  Section 7.3.3.2.

            7.4.2  Method  8000 provides instructions on the analysis sequence,
      appropriate  dilutions,  establishing  daily  retention  time windows,  and
      identification criteria.  Include a mid-concentration standard after each
      group of 10  samples in the analysis sequence.

            7.4.3  Table 1 summarizes the estimated retention times and detection
      limits for a number  of organic compounds analyzable using this method.  An
      example of  the separation  achieved  by Column  1  is  shown  in  Figure 1.
      Figure 2 shows  an example  of the separation achieved using  Column 2.

            7.4.4  Record the sample volume purged or injected and the resulting
      peak sizes  (in  area units  or peak heights).

            7.4.5  Calculation of concentration is covered in Method 8000.
                                   8020A -  6                         Revision 1
                                                                September 1994

-------
            7.4.6  If analytical  interferences are suspected, or for the purpose
      of confirmation, analysis using the second GC column is recommended.

            7.4.7  If the response  for  a peak is off  scale,  i.e.,  beyond the
      calibration range of the standards, prepare a dilution of the sample with
      organic-free reagent water.   The  dilution must  be performed on a second
      aliquot of the sample which has been properly sealed and stored prior to
      use.
8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One for specific quality  control  procedures and
Method 8000 for gas chromatographic procedures.  Quality control to ensure the
proper operation of the purge-and-trap device is covered in Method 5030.

      8.2   Quality control required to validate the GC system operation is found
in Method 8000.

            8.2.1  The  quality control check  sample  concentrate (Method 8000)
      should contain each parameter of interest  at  a concentration of 10 mg/L
      in methanol.

            8.2.2  Table 3  indicates the calibration  and QC acceptance criteria
      for this method.   Table 4 gives method accuracy  and precision as functions
      of concentration  for the  analytes  of interest.   The contents of both
      tables should be  used to evaluate a  laboratory's  ability  to perform and
      generate acceptable data by this method.

      8.3   Calculate surrogate  standard recovery on  all  samples,  blanks, and
spikes.   Determine if recovery  is  within limits (limits established by performing
QC procedure outlined in Method 8000).

            8.3.1  If recovery is not  within limits, the following is required.

                   •    Check   to  be  sure  that  there   are   no   errors  in
                        calculations,    surrogate   solutions   and   internal
                        standards.  Also,  check instrument  performance.

                   •    Recalculate the data and/or reanalyze the extract if any
                        of  the above  checks reveal  a  problem.

                   •    Reextract and  reanalyze the sample  if none of the above
                        are   a   problem  or   flag  the  data   as   "estimated
                        concentration".
                                  8020A - 7                         Revision 1
                                                                September 1994

-------
9.0   METHOD PERFORMANCE

      9.1   This method was tested by 20 laboratories  using organic-free reagent
water, drinking water,  surface water, and three industrial wastewaters spiked at
six concentrations over the range 2.1  -  500  jiig/L.   Single operator precision,
overall precision, and method accuracy were found to  be directly related to the
concentration of the  parameter and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 4.

      9.2   The accuracy and precision obtained will be determined  by the sample
matrix, sample introduction technique, and by the calibration procedure used.

      9.3   The method detection  limits reported in Table  1 were generated under
optimum analytical conditions by  an  Agency contractor (Ref.  7) as guidance, and
may not be readily achievable by all laboratories at all  times.


10.0  REFERENCES

1.    Bellar, T.A., and J.J.  Lichtenberg, J.  Amer.  Water Works Assoc.,  66(12),
      pp. 739-744, 1974.

2.    Bellar, T.A., and J.J.  Lichtenberg,  "Semi-Automated Headspace Analysis of
      Drinking  Waters  and Industrial  Waters  for  Purgeable  Volatile  Organic
      Compounds", in  Van Hall  (ed.),  Measurement of Organic  Pollutants in Water
      and Wastewater, ASTM STP 686,  pp. 108-129,  1979.

3.    Dowty, B.J., S.R.  Antoine, and J.L.  Laseter, "Quantitative and Qualitative
      Analysis of Purgeable Organics by High  Resolution Gas Chromatography and
      Flame  lonization  Detection",  in Van Hall,  ed., Measurement  of  Organic
      Pollutants in Water and Wastewater.  ASTM STP 686,  pp. 24-35, 1979.

4.    Development and Application of Test  Procedures  for Specific Organic Toxic
      Substances  in  Wastewaters.   Category 11 - Purgeables  and  Category 12 -
      Acrolein,  Acrylonitrile,  and Dichlorodifluoromethane.   Report  for EPA
      Contract 68-03-2635.

5.    "EPA Method Validation Study 24, Method  602  (Purgeable Aromatics)", report
      for EPA Contract 68-03-2856.

6.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean  Water Act;  Final Rule and Interim
      Final Rule and Proposed Rule", October  26,  1984.

7.    Gebhart, J.E.,  S.V. Lucas,  S.J. Naber,  A.M. Berry,  T.H. Danison and H.M.
      Burkholder,  "Validation of  SW-846  Methods  8010,  8015, and 8020"; Report
      for  EPA  Contract 68-03-1760,  Work Assignment 2-15;    US EPA,  EMSL-
      Cincinnati, 1987."
                                   8020A -  8                        Revision 1
                                                                September 1994

-------
                                  TABLE 1.
            CHROMATOGRAPHIC  CONDITIONS  AND METHOD  DETECTION  LIMITS
                        FOR AROMATIC VOLATILE ORGANICS



Compound
Benzene
Chlorobenzeneb
1,4-Dichlorobenzene
1,3-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl Benzene
Toluene
Xylenes
Retention
(min)

Col. 1
3.33
9.17
16.8
18.2
25.9
8.25
5.75

time


Col. 2
2.75
8.02
1'6. 2
15.0
19.4
6.25
4.25

Method
detection
limit3
(M9/L)
0.2
0.2
0.3
0.4
0.4
0.2
0.2

a Using purge-and-trap method (Method 5030). See Sec. 9.3.
b Chlorobenzene and m-xylene may co-elute on some columns.
                                  TABLE 2.
             DETERMINATION  OF  ESTIMATED  QUANTITATION  LIMITS  (EQLs)
                             FOR VARIOUS MATRICES3
      Matrix
Factor
      Ground water
      Low-concentration soil
      Water miscible liquid waste
      High-concentration soil and sludge
      Non-water miscible waste
  10
  10
 500
1250
1250
      EQL =  [Method detection limit  (see  Table 1)]  X [Factor found  in  this
      table].  For  non-aqueous  samples,  the factor is on  a wet-weight basis.
      Sample EQLs are highly matrix-dependent.   The EQLs determined herein are
      provided for guidance and may not always  be achievable.
                                   8020A  -  9
              Revision 1
          September 1994

-------
                             TABLE 3.
                      QC ACCEPTANCE CRITERIA3


Parameter
Benzene
Chlorobenzene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Di chlorobenzene
Ethyl benzene
Toluene
Range
for Q
(M9/L)
15.4-24.6
16.1-23.9
13.6-26.4
14.5-25.5
13.9-26.1
12.6-27.4
15.5-24.5
Limit
for s
(M9/L)
4.1
3.5
5.8
5.0
5.5
6.7
4.0
Range
for x
(M9/L)
10.0-27.9
12.7-25.4
10.6-27.6
12.8-25.5
11.6-25.5
10.0-28.2
11.2-27.7
Range
P> Ps
(%)
39-150
55-135
37-154
50-141
42-143
32-160
46-148
Q     =
s     =
x     =
P, Ps =
      Concentration measured in QC check sample, in
      Standard deviation of four recovery measurements, in M9/L-
      Average recovery for four recovery measurements, in jug/L.
      Percent recovery measured.
Criteria from 40 CFR  Part  136  for Method 602,  using packed columns, and
were calculated assuming a check sample concentration of 20 fj.g/1.  These
criteria are based directly upon the method performance data in Table 4.
Where necessary, the  limits  for recovery have  been  broadened  to assure
applicability of the limits to  concentrations below those used to develop
Table 1. When capillary columns are used, see Method  8021 for performance
data.
                            8020A - 10
                                                                    Revision 1
                                                                September 1994

-------
                                   TABLE 4.
          METHOD ACCURACY AND PRECISION AS  FUNCTIONS OF CONCENTRATION


Parameter
Benzene
Chlorobenzene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Di chlorobenzene
Ethyl benzene
Toluene
Accuracy, as
recovery, x'
(M9/L)
0.92C+0.57
0.95C+0.02
0.93C+0.52
0.96C-0.04
0.93C-0.09
0.94C+0.31
0.94C+0.65
Single analyst
precision, sr'
(M9/L)
0.09x+0.59
0.09X+0.23
0.17X-0.04
0.15x-0.10
0.15X+0.28
0.17X+0.46
0.09X+0.48
Overall
precision,
S' (M9/L)
0.21X+0.56
0.17X+0.10
0.22X+0.53
0.19X+0.09
0.20X+0.41
0.26X+0.23
0.18x+0.71
x'
Expected  recovery  for  one  or  more  measurements   of  a  sample
containing concentration C, in
            Expected single  analyst  standard deviation of  measurements at an
            average concentration of x,  in
S'


C

x
Expected inter! aboratory standajrd  deviation  of measurements at an
average concentration found of x, in
True value for the concentration, in /^g/L.

Average recovery  found for  measurements  of  samples  containing a
concentration of C, in M9/L.
                                  8020A - 11
                                                        Revision 1
                                                    September 1994

-------
                  Figure 1
Chromatogram of Aromatic Volatile Organics
           (column 1 conditions)

                   Column:
                   Program:
                   Detector:
                   Sample:
5% SP-1200/1.75% Bentone 34
50°C-2 Minutes, 6°C/Min. to 90°C
Photoionization
0.40 jig/L Standard Mixture
         8       10      12      14

          RETENTION TIME (MINUTES}
        16
18
20      22
                 8020A - 12
                    Revision 1
                September 1994

-------
                  Figure 2
Chromatogram of Aromatic Volatile Organics
           (column 2 conditions)
Column:
Program:
Detector:
Sample:
  5% l,2,3-Tris(2-Cyanoethoxy)Propane on Chromosorb-W
  40°C-2 Minutes, 2°C/Min. to 100*C
  Photoionization
  2.0 /jg/L Standard Mixture
 I

mm
                    12       If

                        (MINUTU>
24
                 8020A - 13
                                         Revision  1
                                     September 1994

-------
                                     METHOD  8020A
               AROMATIC  VOLATILE ORGANICS BY  GAS  CHROMATOGRAPHY
        Start
7.1  Introduce compounds
into gas chromatograph
  by direct injection or
    purge-and-trap
    (Method 5030)
      7.2 Set gas
    chromatograph
      condition.
     7.3 Calibrate
 (refer to Method 8000)
    7.4.1 Introduce
  volatile compounds
into gas chromatograph
  by purge-and-trap or
    direct injection.
  7.4.2 Follow Method
   8000 for analysis
    sequence, etc.
7.4.4 Record volume
 purged or injected
  and peak sizes.
    7.4.5 Calculate
    concentration
(refer to Method 8000)
      7.4.6 Are
      analytical
    interferences
     suspected?
                                            7.4.7 Is
                                          response for
                                             a peak
                                           off-scale?
                                                                   7.4.6 Analyze using
                                                                    second GC column.
                               7.4.7 Dilute second
                                aliquot of sample.
                                       8020A -14
                                            Revision  1
                                       September  1994

-------
                                 METHOD 8021A

               HALOGENATED VOLATILES  BY  GAS CHROMATOGRAPHY  USING
            PHOTOIONIZATION AND ELECTROLYTIC  CONDUCTIVITY DETECTORS
                    IN SERIES:  CAPILLARY  COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8021  is  used  to determine volatile organic  compounds  in a
variety of solid waste matrices.   This method  is applicable to nearly all types
of samples,  regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily wastes,  mousses,  tars,
fibrous  wastes,  polymeric  emulsions,   filter  cakes,  spent  carbons,  spent
catalysts, soils, and sediments.   The following compounds can be determined by
this method:
Analyte
                                                  Appropriate Technique
CAS No.a    Purge-and-Trap
Direct
Injection
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chl orodi bromomethane
Chloroethane
Chloroform
Chloromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Di bromomethane
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1, 4 -Di chl orobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1, 1-Dichloroethene
cis-1, 2-Di chl oroethene
trans -1, 2-Di chl oroethene
71-43-2
108-86-1
74-97-5
75-27-4
75-25-2
74-83-9
104-51-8
135-98-8
98-06-6
56-23-5
108-90-7
124-48-1
75-00-3
67-66-3
74-87-3
95-49-8
106-43-4
96-12-8
106-93-4
74-95-3
95-50-1
541-73-1
106-46-7
75-71-8
75-34-3
107-06-2
75-35-4
156-59-4
156-60-5
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
PP
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
                                  8021A  -  1
                               Revision 1
                           September 1994

-------
                                                  Appropriate Technique
Analyte
CAS No.a    Purge-and-Trap
Direct
Injection
1 , 2-Di chl oropropane
1,3-Dichloropropane
2, 2-Di chl oropropane
1,1-Dichloropropene
cis-l,3-dichloropropene
trans-l,3-dichloropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p-Isopropyl toluene
Methylene chloride
Naphthalene
n-Propyl benzene
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1-Tri chl oroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chl orof 1 uoromethane
1 , 2 , 3 -Tri chl oropropane
1, 2, 4-Tri methyl benzene
1,3, 5 -Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
78-87-5
142-28-9
590-20-7
563-58-6
10061-01-5
10061-02-6
100-41-4
87-68-3
98-82-8
99-87-6
75-09-2
91-20-3
103-65-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
87-61-6
120-82-1
71-55-6
79-00-5
79-01-6
75-69-4
96-18-4
95-63-6
108-67-8
75-01-4
95-47-6
108-38-3
106-42-3
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
b
a Chemical Abstract Services Registry Number.
b Adequate response by thi
pp Poor purging efficiency
s technique.
resulting in high EQLs.




                                       is
       1.2    Method detection limits (MDLs) are compound dependent  and vary with
purging  efficiency  and  concentration.    The MDLs  for selected  analytes  are
presented  in  Table  1.   The  applicable  concentration  range  of  this method
compound  and  instrument dependent  but  is   approximately  0.1  to 200
Analytes that  are inefficiently purged  from water will not  be detected when
present at low concentrations,  but they can be measured with acceptable accuracy
and  precision when  present  in sufficient  amounts.   Determination  of some
structural isomers (i.e. xylenes) may be hampered by coelution.
                                   8021A -  2
                               Revision 1
                           September 1994

-------
      1.3    The  estimated  quantitation  limit  (EQL)  of  Method  8021A  for an
individual  compound  is approximately 1  //g/kg (wet weight)  for  soil/sediment
samples, 0.1 mg/kg  (wet  weight)  for wastes, and 1 /zg/L  for  ground water  (see
Table 3).   EQLs will be proportionately higher for sample extracts and samples
that require dilution or reduced sample size  to avoid saturation of the detector.

      1.4    This method  is  recommended for  use only by analysts experienced in
the measurement of purgeable organics at the  low  /Ltg/L level  or by experienced
technicians under the close supervision of a qualified analyst.

      1.5    The toxicity or carcinogenicity  of chemicals used in this method has
not been  precisely  defined.  Each  chemical  should be treated as  a potential
health  hazard,  and exposure  to these  chemicals   should  be  minimized.   Each
laboratory is responsible for maintaining  awareness of OSHA regulations regarding
safe  handling  of chemicals used in  this   method.   Additional  references to
laboratory  safety are available for the information of the analyst (references
4 and 6).

      1.6    The  following  method  analytes  have  been tentatively  classified as
known or suspected human or mammalian carcinogens: benzene,  carbon tetrachloride,
1,4-dichlorobenzene,    1,2-dichloroethane,    hexachloro-butadiene,    1,1,2,2-
tetrachloroethane,    1,1,2-trichloroethane,    chloroform,   1,2-dibromoethane,
tetrachloroethene,  trichloroethene,  and vinyl chloride.  Pure standard materials
and stock standard  solutions of these compounds  should  be  handled in a hood.  A
NIOSH/MESA approved toxic gas respirator should be worn when the analyst handles
high concentrations of these toxic compounds.


2.0   SUMMARY OF METHOD

      2.1    Method  8021   provides  gas  chromatographic  conditions  for  the
detection of halogenated and aromatic volatile organic compounds.  Samples can
be analyzed using direct injection or purge-and-trap (Method 5030). Ground water
samples must be analyzed  using Method 5030  (where applicable).   A temperature
program  is  used  in the  gas chromatograph  to separate  the  organic compounds.
Detection is achieved  by  a  photoionization  detector  (PID) and an electrolytic
conductivity detector  (HECD) in series.

      2.2    Tentative  identifications are obtained by analyzing standards under
the same conditions  used  for samples and  comparing  resultant GC retention times.
Confirmatory information can be gained by comparing the relative response from
the two detectors.   Concentrations of the identified components are measured by
relating the response produced for that compound to the response produced by a
compound that is used as an internal standard.


3.0   INTERFERENCES

      3.1    Refer to Methods 5030 and 8000.

      3.2    Samples  can  be contaminated  by diffusion  of volatile  organics
(particularly chlorofluorocarbons and  methylene chloride) through  the  sample
container  septum  during  shipment  and storage.   A trip  blank  prepared  from


                                  8021A  - 3                        Revision 1
                                                                September 1994

-------
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.

      3.3    Sulfur dioxide is a potential  interferant  in the analysis for vinyl
chloride.
4.0   APPARATUS AND MATERIALS

      4.1    Sample  introduction  apparatus  -  Refer to  Method  5030 for  the
appropriate equipment for sample introduction purposes.

      4.2    Gas  Chromatograph  - capable of temperature  programming; equipped
with variable-constant differential flow controllers, subambient oven controller,
photoionization and electrolytic conductivity detectors connected with a short
piece of uncoated capillary tubing, 0.32-0.5 mm ID, and data system.

             4.2.1  Column  -  60  m x 0.75 mm ID VOCOL wide-bore capillary column
      with 1.5 nm  film  thickness  (Supelco Inc., or  equivalent).

             4.2.2  Photoionization   detector  (PID)  (Tracer   Model   703,   or
      equivalent).

             4.2.3  Electrolytic conductivity detector (HECD)  (Tracor Hall Model
      700-A, or equivalent).

      4.3    Syringes -  5  ml glass  hypodermic with  Luer-Lok  tips.

      4.4    Syringe valves  - 2-way with  Luer ends  (Teflon or Kel-F).

      4.5    Microsyringe  -  25  juL  with  a  2 in. x 0.006  in. ID,  22° bevel needle
(Hamilton #702N or equivalent).

      4.6    Microsyringes - 10, 100 juL.

      4.7    Syringes -  0.5, 1.0,  and 5 ml, gas-tight with shut-off  valve.

      4.8    Bottles - 15  ml, Teflon lined with screw-cap or crimp top.

      4.9    Analytical  balance - 0.0001  g.

      4.10   Refrigerator.

      4.11   Volumetric  flasks, Class  A  - Appropriate  sizes with ground glass
stoppers.


5.0   REAGENTS

      5.1    Reagent  grade  inorganic  chemicals shall  be used in  all  tests.
Unless otherwise  indicated,  it is intended that all  inorganic reagents shall
conform  to the  specifications  of the Committee on  Analytical  Reagents  of the
American Chemical  Society, where such specifications  are available. Other grades


                                   8021A  - 4                         Revision 1
                                                                September 1994

-------
may be used, provided it is first  ascertained that the reagent  is of sufficiently
high  purity  to  permit  its  use  without  lessening  the  accuracy  of  the
determination.

      5.2   Organic-free  reagent water.  All  references to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Methanol, CH3OH - Pesticide quality or equivalent, demonstrated to
be free of analytes.  Store away from other solvents.

      5.4   Vinyl chloride, (99.9% pure), CH2=CHC1.  Vinyl chloride is available
from  Ideal  Gas  Products,  Inc.,  Edison,  New Jersey  and  from Matheson,  East
Rutherford, New Jersey, as well  as  from other  sources.  Certified mixtures of
vinyl chloride in nitrogen at  1.0 and  10.0  ppm  (v/v)  are available from several
sources.

      5.5   Stock standards - Stock solutions may either be prepared from pure
standard materials or purchased as certified solutions.  Prepare stock standards
in methanol using assayed liquids or  gases,  as appropriate.  Because  of the
toxicity of some of the organohalides, primary dilutions of these materials of
the toxicity should  be prepared in a hood.

      NOTE: If  direct  injection  is  used,  the solvent system of standards must
            match that  of the  sample.   It  is  not  necessary to  prepare high
            concentration aqueous mixed standards when using direct injection.

            5.5.1 Place about 9.8 ml  of methanol  in  a  10 mL tared ground glass
      stoppered  volumetric flask.   Allow  the flask  to stand, unstoppered, for
      about 10 minutes until  all  alcohol-wetted surfaces have dried.  Weigh the
      flask to the nearest 0.1 mg.

            5.5.2 Add  the assayed  reference  material,  as described  below.

                  5.5.2.1    Liquids:  Using a  100 jiL  syringe, immediately add
            two  or more drops of assayed reference material to the flask; then
            reweigh.   The liquid must fall  directly  into  the  alcohol  without
            contacting the neck  of  the  flask.

                  5.5.2.2    Gases:   To  prepare standards  for  any compounds
            that   boil   below   30°C   (e.g.    bromomethane,    chloroethane,
            chloromethane,  dichlorodifluoromethane,   trichlorofluoromethane,
            vinyl chloride),  fill  a  5 ml valved gas-tight syringe with the
            reference standard  to  the 5.0 ml mark.    Lower the needle  to 5 mm
            above  the  methanol  meniscus.    Slowly  introduce  the  reference
            standard above the  surface of the  liquid.  The  heavy gas  rapidly
            dissolves in  the methanol.  This may  also  be accomplished by using
            a  lecture  bottle  equipped  with  a  Hamilton Lecture  Bottle  Septum
            (#86600).  Attach  Teflon  tubing  to the  side-arm  relief valve and
            direct a gentle stream of gas  into  the methanol meniscus.

            5.5.3 Reweigh, dilute to volume,  stopper, and then mix by inverting
      the flask  several times.   Calculate  the  concentration  in  milligrams per
      liter (mg/L) from the net gain in weight.   When  compound purity is assayed


                                  8021A -  5                         Revision 1
                                                                September 1994

-------
      to  be 96%  or greater,  the  weight  may  be  used  without  correction  to
      calculate the concentration of the stock  standard.  Commercially prepared
      stock standards may be used at any concentration if they are certified by
      the manufacturer or by an  independent source.

            5.5.4  Transfer  the  stock  standard  solution  into a  bottle  with a
      Teflon lined  screw-cap or  crimp  top.   Store,  with  minimal  headspace,  at
      -10°C to -20°C and  protect  from light.

            5.5.5  Prepare  fresh  stock  standards for gases  weekly or sooner if
      comparison with check standards indicates a problem.   Reactive compounds
      such as 2-chloroethyl  vinyl ether  and styrene may need to be prepared more
      frequently.  All other standards  must be  replaced after six months.  Both
      gas and liquid  standards must  be  monitored  closely  by comparison to the
      initial calibration curve and by comparison to QC check standards.  It may
      be  necessary to replace the  standards  more frequently if  either check
      exceeds a 20% drift.

      5.6   Prepare   secondary  dilution  standards,  using   stock  standard
solutions, in methanol, as needed, that contain the compounds of interest, either
singly or mixed together.  The secondary dilution standards should be prepared
at concentrations such that the  aqueous calibration standards prepared in Sec.
5.7 will  bracket the working range of the analytical  system.  Secondary dilution
standards should be stored with  minimal  headspace  for volatiles  and should be
checked frequently for signs of degradation or evaporation, especially just prior
to preparing calibration standards from them.

      5.7   Calibration standards, at a  minimum of five concentration levels are
prepared in organic-free reagent  water  from the  secondary dilution of the stock
standards.  One of the concentration levels should be at  a concentration near,
but above, the method detection limit.  The remaining concentration levels should
correspond to the  expected range  of the  concentrations found in real samples or
should define  the working range of the GC.   Standards  (one or  more)  should
contain each analyte for  detection  by this method.   In order to prepare accurate
aqueous standard solutions, the  following precautions must be observed.

      NOTE: Prepare calibration solutions for use with direct injection analyses
            in water  at  the concentrations  required.

            5.7.1  Do  not inject more  than  20  /^L  of  alcoholic  standards into
      100 mL of water.

            5.7.2  Use  a  25   juL  Hamilton  702N  microsyringe  or  equivalent
      (variations  in needle  geometry   will  adversely affect  the  ability  to
      deliver reproducible volumes of methanolic standards  into water).

            5.7.3  Rapidly   inject   the   alcoholic  standard  into   the  filled
      volumetric flask.  Remove the needle  as fast  as possible after injection.

            5.7.4  Mix aqueous  standards by inverting  the flask three  times.
                                   8021A -  6                        Revision 1
                                                                September 1994

-------
             5.7.5 Fill  the sample syringe  from the standard solution contained
       in  the  expanded  area of the flask (do not  use any solution contained in
       the  neck  of the flask).

             5.7.6 Never use  pipets  to  dilute or transfer  samples  or aqueous
       standards.

             5.7.7 Aqueous standards are not stable and should be discarded after
       one  hour,  unless  properly sealed  and stored.   The aqueous standards can
       be  stored  up to 12  hours,  if held  in  sealed vials with zero headspace.

             5.7.8 Optionally calibration using a certified gaseous mixture can
       be  accomplished  daily  utilizing  commercially  available  gaseous analyte
       mixture  of  bromomethane,  chloromethane,  chloroethane,  vinyl  chloride,
       dichlorodifluoromethane and trichlorofluoromethane  in  nitrogen.  These
       mixtures  of  documented  quality are  stable  for as  long as  six months
       without refrigeration.  (VOA-CYL III,  RESTEK  Corporation, Cat.  #20194 or
       equivalent).

       5.8    Internal   standards  -   Prepare  a  spiking  solution  containing
fluorobenzene and  2-bromo-l-chloropropane  in  methanol,  using  the  procedures
described  in Sees.  5.5 and 5.6.   It is recommended that the secondary dilution
standard  be  prepared  at a concentration of 5 mg/L  of  each  internal  standard
compound.   The  addition of  10  nl of such  a  standard to 5.0  ml  of  sample or
calibration standard would be equivalent to 10 M9/L.

       5.9    Surrogate   standards  -   The  analyst   should  monitor  both  the
performance of  the analytical  system and  the effectiveness of the  method in
dealing with each  sample  matrix by  spiking each  sample,  standard,  and reagent
blank with two or more surrogate compounds.   A  combination of 1,4-dichlorobutane
and bromochlorobenzene is recommended to encompass the range of the temperature
program used in  this method.   From stock standard solutions prepared as in Sec.
5.5, add  a volume to give 750  /^g of each  surrogate  to  45  mL  of organic-free
reagent water contained in a  50 ml volumetric flask, mix, and dilute to volume
for a concentration of 15 ng//xL.  Add 10 p,l of this surrogate spiking solution
directly into the 5 mL syringe with every sample and reference standard analyzed.
If the internal  standard calibration  procedure  is used, the surrogate compounds
may be added directly to the  internal standard spiking solution (Sec. 5.8).


6.0    SAMPLE COLLECTION,  PRESERVATION, AND  HANDLING

       6.1   See  the introductory material   to this  chapter,  Organic  Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into  the gas chromatograph either
by direct injection or purge-and-trap  (Method  5030).  Method  5030 may be used
directly on ground  water samples or low-concentration  contaminated  soils and
sediments.   For medium-concentration  soils or sediments,  methanolic extraction,
as described in Method  5030,  may  be necessary prior to purge-and-trap analysis.


                                  8021A  - 7                         Revision 1
                                                                September 1994

-------
      7.2   Gas chromatography conditions  (Recommended)

            7.2.1  Set   up   the   gas   chromatograph  system   so   that   the
      photoionization  detector  (PID)   is   in  series  with  the  electrolytic
      conductivity detector  (HECD).

            7.2.2  Oven  settings:

            Carrier gas  (Helium) Flow rate:     6 mL/min.
            Temperature  program
                   Initial  temperature:     10°C,  hold for 8 minutes at
                   Program:                 10°C to 180°C  at 4°C/"un
                   Final  temperature:       180°C,   hold   until    all   expected
                                           compounds have  eluted.

            7.2.3  The carrier gas flow is augmented  with  an additional 24 ml of
      helium flow  before  entering  the  photoionization  detector.   This make-up
      gas is necessary to ensure optimal response from both detectors.

            7.2.4  These  halogen-specific systems eliminate misidentifications
      due to non-organohalides which are coextracted during the purge step.  A
      Tracer Hall  Model  700-A detector  was  used to gather  the single laboratory
      accuracy  and  precision data  presented   in  Table  2.    The  operating
      conditions used to collect these data are:

            Reactor tube:                       Nickel, 1/16 in OD
            Reactor temperature:                810°C
            Reactor base  temperature:           250°C
            Electrolyte:                        100% n-Propyl  alcohol
            Electrolyte  flow rate:              0.8 mL/min
            Reaction gas:                       Hydrogen at 40 mL/min
            Carrier gas  plus make-up gas:       Helium at 30 mL/min

            7.2.5  A  sample chromatogram  obtained with this column is presented
      in  Figure  5.  This column was  used to develop the method performance
      statements in Sec.  9.0.  Estimated retention  times and  MDLs that can be
      achieved under these conditions  are  given  in  Table 1.   Other columns or
      element specific detectors  may be used if the requirements of Sec. 8.0 are
      met.

      7.3   Calibration   -   Refer  to   Method  8000  for  proper  calibration
techniques.   Use Table  1  and especially  Table  2  for guidance  on selecting the
lowest point on the calibration curve.

            7.3.1  Calibration must take place using the same sample introduction
      method that will be used to analyze  actual samples  (see Sec. 7.4.1).

            7.3.2  The  procedure  for internal  or external calibration  may be
      used.  Refer to Method  8000 for a description of each of these procedures.

      7.4   Gas chromatographic  analysis
                                   8021A -  8                        Revision 1
                                                                September 1994

-------
             7.4.1  Introduce volatile compounds into the gas chromatograph using
      either Method 5030 (purge-and-trap method)  or the direct injection method
      (see  Sec.  7.4.1.1).   If the internal  standard  calibration  technique is
      used, add  10 /zL  of  internal  standard to the  sample prior to  purging.

                   7.4.1.1     Direct injection - In very  limited  applications
             (e.g.  aqueous  process  wastes)  direct injection  of the sample into
             the  GC  system with a 10  /zL  syringe  may  be  appropriate.   The
             detection limit is very high (approximately 10,000 M9/L)> therefore,
             it is  only permitted where concentrations in excess of 10,000 M9A
             are  expected or for water-soluble compounds  that do not purge.  The
             system must be calibrated by direct injection (bypassing the purge-
             and-trap device).

             7.4.2  Follow   Sec.  7.6  in  Method 8000 for  instructions  on  the
      analysis sequence,  appropriate dilutions,  establishing  daily  retention
      time  windows,  and  identification  criteria.  Include  a mid-concentration
      standard after each  group of 10 samples in the analysis sequence.

             7.4.3  Table  1  summarizes the estimated retention  times on the two
      detectors  for a number of organic  compounds analyzable using this method.

             7.4.4  Record the  sample volume  purged or injected and the resulting
      peak  sizes (in area  units or peak heights).

             7.4.5  Calculation of concentration is  covered  in Method 8000.

             7.4.6  If analytical interferences are suspected, or for the purpose
      of confirmation, analysis using a second GC column is recommended.

             7.4.7  If the   response for  a peak is off-scale,  i.e.,  beyond the
      calibration range of the standards, prepare a  dilution of the sample with
      organic-free reagent water.  The  dilution  must  be  performed on  a second
      aliquot of the sample which  has been  properly sealed and stored prior to
      use.
8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter One for specific quality  control  procedures  and
Method 8000 for gas chromatographic procedures.   Quality control to ensure  the
proper operation of the purge-and-trap device is  covered in Method 5030.

      8.2   Quality control  required to  validate  the GC  system  operation  is
found in Method 8000.

            8.2.1  The  quality control  reference  sample (Method  8000)  should
      contain  each parameter  of interest at  a  concentration  of 10 mg/L  in
      methanol.

            8.2.2  Table  2  gives  method  accuracy  and  precision as  functions  of
      concentration for the analytes of interest.
                                  8021A  - 9                         Revision 1
                                                                September 1994

-------
      8.3    Calculate  surrogate  standard  recovery on all  samples,  blanks, and
spikes.  Determine  if recovery is  within limits (limits established by performing
QC procedure outlined in Method 8000).

             8.3.1  If recovery is not  within limits,  the following is required.

                   •     Check to be sure there  are  no errors in calculations,
                        surrogate solutions  and  internal standards.  Also check
                        instrument  performance.

                   •     Recalculate the data and/or reanalyze the extract  if any
                        of the above checks  reveal  a problem.

                   •     Reextract and  reanalyze  the  sample if none of the above
                        are  a   problem   or   flag   the  data   as   "estimated
                        concentration".
9.0   METHOD PERFORMANCE

      9.1   Method detection limits for these analytes have been calculated from
data collected by spiking organic-free reagent water  at  0.1  /ng/L.   These data
are presented in Table 1.

      9.2   This  method  was tested in a  single  laboratory using organic-free
reagent water spiked at 10 jug/L.   Single laboratory precision and accuracy data
for each detector are presented for the method analytes in Table 2.


10.0  REFERENCES

1.    Volatile Organic Compounds  in Water by Purqe-and-Trap Capillary Column Gas
      Chromatography   with   Photoionization  and   Electrolytic   Conductivity
      Detectors  in  Series,  Method 502.2.  Rev.  2.0  (1989);  Methods  for  the
      Determination  of Organic  Compounds  in  Drinking  Water",  Environmental
      Monitoring Systems Laboratory, Cincinnati,  OH, EPA/600/4-88/039, December,
      1988

2.    The Determination of Halogenated Chemicals  in Water by the Purge and Trap
      Method,  Method  502.1;  Environmental  Protection  Agency,  Environmental
      Monitoring  and  Support  Laboratory:  Cincinnati, Ohio  45268,  September,
      1986.

3.    Volatile Aromatic and Unsaturated Organic  Compounds in Water by Purge and
      Trap Gas Chromatography, Method 503.1;  Environmental Protection Agency,
      Environmental  Monitoring  and  Support  Laboratory:  Cincinnati,  Ohio,
      September,  1986.

4.    Glaser, J.A.; Forest, D.L.; McKee, G.D.; Quave, S.A.; Budde,  W.L. "Trace
      Analyses for Wastewaters"; Environ. Sci. Technol.  1981, 15, 1426.
                                  8021A - 10                        Revision 1
                                                                September 1994

-------
5.     Bellar, T.A.;  Lichtenberg,  J.J.  The Determination  of Synthetic Organic
      Compounds in Water by Purge and Sequential Trapping Capillary Column Gas
      Chromatography;  U.S.  Environmental  Protection  Agency,  Environmental
      Monitoring and Support Laboratory: Cincinnati, Ohio, 45268.
                                  8021A - 11                        Revision 1
                                                                September 1994

-------
                              TABLE  1.

CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL) FOR
  VOLATILE ORGANIC COMPOUNDS ON PHOTOIONIZATION DETECTION (PID) AND
       HALL  ELECTROLYTIC  CONDUCTIVITY  DETECTOR  (HECD) DETECTORS
Analyte
Di chl orod i f 1 uoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orof 1 uoromethane
1,1-Dichloroethene
Methylene Chloride
trans-l,2-Dichloroethene
1,1-Dichl oroethane
2,2-Dichloropropane
cis-l,2-Di chloroethane
Chloroform
Bromochl oromethane
1,1,1 -Tri chl oroethane
1 , 1-Dichl oropropene
Carbon Tetrachloride
Benzene
1,2-Dichloroethane
Trichloroethene
1,2-Dichloropropane
Bromodi chl oromethane
Dibromomethane
Toluene
1,1, 2 -Tri chl oroethane
Tetrachl oroethene
1,3-Dichloropropane
Di bromochl oromethane
1,2-Dibromoethane
Chlorobenzene
Ethyl benzene
1,1,1 , 2-Tetrachl oroethane
m-Xylene
p-Xylene
o-Xylene
Styrene
Isopropyl benzene
Bromoform
1,1,2 , 2-Tetrachl oroethane
1,2,3-Trichloropropane
PID
Ret. Time8
minute
_b
-
9.88
-
-
-
16.14
-
19.30
-
-
23.11
-
-
-
25.21
-
26.10
-
27.99
-
-
-
31.95
-
33.88
-
-
-
36.56
36.72
-
36.98
36.98
38.39
38.57
39.58
-
-
-
HECD
Ret. Time
minute
8.47
9.47
9.93
11.95
12.37
13.49
16.18
18.39
19.33
20.99
22.88
23.14
23.64
24.16
24.77
25.24
25.47
-
26.27
28.02
28.66
29.43
29.59
-
33.21
33.90
34.00
34.73
35.34
36.59
-
36.80
-
-
-
-
-
39.75
40.35
40.81
PID
MDL
M9/L


0.02



ND°

0.05


0.02



0.02

0.009

0.02



0.01

0.05



0.003
0.005

0.01
0.01
0.02
0.01
0.05



HECD
MDL
M9/L
0.05
0.03
0.04
1.1
0.1
0.03
0.07
0.02
0.06
0.07
0.05
0.01
0.02
0.01
0.03
0.02
0.01

0.03
0.01
0.006
0.02
2.2

ND
0.04
0.03
0.03
0.8
0.01

0.005





1.6
0.01
0.4
                              8021A -  12
    Revision 1
September 1994

-------
                                   TABLE 1.
                                  (Continued)
Analyte
   PID
Ret. Time8
  minute
  HECD
Ret. Time
 minute
                                                               PID
                                                               MDL
HECD
 MDL
n-Propylbenzene                   40.87
Bromobenzene                      40.99
1,3,5-Trimethylbenzene            41.41
2-Chlorotoluene                   41.41
4-Chlorotoluene                   41.60
tert-Butylbenzene                 42.92
1,2,4-Trimethylbenzene            42.71
sec-Butyl benzene                  43.31
p-Isopropyltoluene                43.81
1,3-Dichlorobenzene               44.08
1,4-Dichlorobenzene               44.43
n-Butylbenzene                    45.20
1,2-Dichlorobenzene               45.71
l,2-Dibromo-3-Chloropropane
1,2,4-Trichlorobenzene            51.43
Hexachlorobutadiene               51.92
Naphthalene                       52.38
1,2,3-Trichlorobenzene            53.34

Internal Standards
  Fluorobenzene                   26.84
  2-Bromo-1-chloropropane
                 41.03

                 41.45
                 41.63
                 44.11
                 44.47

                 45.74
                 48.57
                 51.46
                 51.96

                 53.37
                33.08
             0.004
             0.006
             0.004
             ND
             0.02
             0.06
             0.05
             0.02
             0.01
             0.02
             0.007
             0.02
             0.05

             0.02
             0.06
             0.06
             ND
0.03

0.01
0.01
0.02
0.01

0.02
3.0
0.03
0.02

0.03
    Retention times determined  on  60 m x  0.75  mm ID  VOCOL  capillary column.
    Program:  Hold at 10°C  for 8 minutes, then program  at 4°C/min to  180°C, and
    hold  until  all  expected compounds have  eluted.
b   Dash  (-)  indicates  detector does  not  respond.

c   ND  =  Not  determined.
                                  8021A - 13
                                  Revision 1
                              September 1994

-------
                   TABLE 2.
SINGLE LABORATORY ACCURACY AND PRECISION DATA
   FOR VOLATILE ORGANIC COMPOUNDS IN WATERd
Photoionization
Detector
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
n-Butylbenzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chl oromethane
2-Chlorotoluene
4-Chlorotoluene
1 ,2-Dibromo-3-chloropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-1,2 Dichloroethene
tr ans-1, 2-Di chl oroethene
1 ,2-Dichloropropane
1,3-Dichloropropane
2,2-Dichloropropane
1 , 1 -Di chl oropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p- 1 sopropyl toluene
Recovery,8
%
99
99
-
-
-
-
100
97
98
-
100
-
-
-
NDC
101
-
-
-
-
102
104
103
-
-
-
100
ND
93
-
-
-
103
101
99
98
98
Standard
Deviation
of Recovery
1.2
1.7
-
-
-
-
4.4
2.6
2.3
-
1.0
-
-
-
ND
1.0
-
-
-
-
2.1
1.7
2.2
-
-
-
2.4
ND
3.7
-
-
-
3.6
1.4
9.5
0.9
2.4
Hall Electrolytic
Conductivity Detector
Standard
Recovery,8 Deviation
% of Recovery
_b
97
96
97
106
97
-
-
-
92
103
96
98
96
97
97
86
102
97
109
100
106
98
89
100
100
103
105
99
103
100
105
103
-
98
-
-
.
2.7
3.0
2.9
5.5
3.7
-
-
-
3.3
3.7
3.8
2.5
8.9
2.6
3.1
9.9
3.3
2.7
7.4
1.5
4.3
2.3
5.9
5.7
3.8
2.9
3.5
3.7
3.8
3.4
3.6
3.4
-
8.3
-
-
                  8021A - 14
    Revision 1
September 1994

-------
                                         TABLE 2.
                                        (Continued)
Analyte
                                  Photoionization
                                      Detector
                                  Recovery,
                           Standard
                           Deviation
                           of Recovery
                                           Hall  Electrolytic
                                           Conductivity Detector
Recovery,
Standard
Deviation
of Recovery
Methyl ene chloride
Naphthalene
n-Propyl benzene
Styrene
1,1, 1 ,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachl oroethene
Toluene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1, 1 -Tri chl oroethane
1 , 1 , 2 -Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1 , 2 , 4 -Tri methyl benzene
1 , 3 , 5 -Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
-
102
103
104
-
-
101
99
106
104
-
-
100
-
-
99
101
109
99
100
99
-
6.3
2.0
1.4
-
-
1.8
0.8
1.9
2.2
-
-
0.78
-
-
1.2
1.4
5.4
0.8
1.4
0.9
97
-
-
-
99
99
97
-
98
102
104
109
96
96
99
-
-
95
-
-
-
2.8
-
-
-
2.3
6.8
2.4
-
3.1
2.1
3.4
6.2
3.5
3.4
2.3
-
-
5.6

-
-
    Recoveries  and standard  deviations  were determined  from  seven samples  and  spiked at
    10  jug/L of each analyte. Recoveries were determined by internal  standard method. Internal
    standards were:  Fluorobenzene  for  PID, 2-Bromo-l-chloropropane for  HECD.
b   Detector  does  not  respond.

c   ND  =  Not  determined.
    This  method was
    reference  5).
tested  in  a  single  laboratory using  water spiked  at 10 M9/L  (see
                                        8021A - 15
                                                     Revision 1
                                                 September 1994

-------
                      TABLE 3.
DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                FOR VARIOUS MATRICES8
  Matrix                              Factor
  Ground water                             10
  Low-concentration soil                   10
  Water miscible liquid waste             500
  High-concentration soil and sludge     1250
  Non-water miscible waste               1250
  EQL = [Method detection limit (see Table 1)] X [Factor found in
  this table].  For non-aqueous samples,  the  factor is on a wet-
  weight basis.  Sample EQLs are highly matrix-dependent.  The EQLs
  listed herein are provided  for  guidance and may  not always be
  achievable.
                     8021A - 16                        Revision 1
                                                   September 1994

-------
                           FIGURE  1.
                        PURGING  DEVICE
                        BUT 114 IN 0.0
                     >— 14 MM 00
                        INLET 1(4 IN. O 0
EXIT 1M IN. 0.0
10 MM GLASS FRTT
SAMPLE INLET


2-WAY SYRINGE VALVE


tr CM 20 GAUGE SYRINGE NEEOU


6 MM 0 0  RUMER SEPTUM


MUTT 1/4 IN 00
                                                      vie IN o o
                                                  y~ STAINLESS STEEL
                                                     13X
                                                     MCXECUCAR SIEVE
                                                     PUflQE GAS FILTER
                                                       PUAGEOAS
                          8021A -  17
                                      Revision  1
                                September 1994

-------
                           FIGURE 2.
TRAP PACKINGS AND CONSTRUCTION TO INCLUDE DESORB  CAPABILITY
       PACKING DETAIL



          ~t- 3 MM GLASS *OX
                                   CONSTRUCTION OCTAJC
             CM SlUO

            'S CM THA* 3C
            — • CM J««. OV-1
                         8021A - 18
    Revision  1
September  1994

-------
                            FIGURE 3.
               PURGE-AND-TRAP  SYSTEM -  PURGE MODE
CAflttEPGAS
PLOW CONTROL
pressure
REGULATOR
                     uomo tfuecnoN PORTS
                           COLUMN OVEN
*
13X MOLECULAR
S46VE FILTER
 I
                                     CONFIRMATORY COLUMN

                                    TO 0€TKTOW

                                     ANALYTICAL COLUMN
                               OPTIONAL 4JSORT COLUMN
                               S€L£CnON VALVf
                                ».	 /- TRAP INLET
                                       TRAP
                                       22"C
                                  PURGING
                                  D€V>C€
                                   NOTE.
                                   ALL UNCS BCTWCEN TRAP
                                   ANO QC SHOULD K HCATEO
                             8021A -  19
                                                   Revision  1
                                               September 1994

-------
                           FIGURE 4.
        SCHEMATIC OF PURGE-AND-TRAP  DEVICE  - DESORB MODE
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE GAS
FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
UOUIO INJECTION PORTS

   r— COLUMN OVEN
                                    JUUV-
                                               CONFIRMATORY COLUMN
                                               TO DETECTOR
                                                ANALYTICAL COLUMN
                              OPTIONAL *PORT COLUMN
                              SELECTION VALVE
                                      TRAP INLET
                                     TRAP
                                     200 *C
                              i PURGING
                               1 DEVICE
             NOTE.
             ALL UNES BETWEEN TRAP
             AND GC SHOULD BE HEATED
             TO«TC.
                           8021A - 20
                            Revision 1
                         September  1994

-------
                        FIGURE 5.

            GAS  CHROMATOGRAM OF VOLATILE ORGAN ICS
COLUMNI 60 METED x 0.73 MM I.D. VOCOL CAPILLARV

    AMI TMAT VOC'l WITH NAU. • flO tH SCAIES

   58 ST  f= R  S3  8CS  8 SSSS538
                                               S^GS £ 8  5 iSSS
JL_JL_L
            x .z  x
                                                 J
                           ILJ
                                            LJOllL.
                                 *- JK I
                                 UN d>-K I
                       8021A - 21
                                                          Revision 1
                                                      September 1994

-------
                                      METHOD  8021A
       HALOGENATED VOLATILES BY  GAS CHROMATOGRAPHY USING PHOTOIONIZATION
                AND  ELECTROLYTIC  CONDUCTIVITY DETECTORS IN  SERIES:
                              CAPILLARY  COLUMN TECHNIQUE
  (     Start     j
      7.2 Set
   chromatographic
     conditions.
    7.3 Refer to
  Method 8000 for
calibration techniques.
  7.4.1  Introduce
sample into GC using
  direct injection or
  purge-and-trap.
    7.4.4 Record
   sample volume
 introduced into GC
   and peak sizes.
   7.4.5 Refer
to Method 8000 for
   calculations.
      7.4.6 Are
      analytical
    interferences
     suspected?
                                      7.4.7 Is peak
                                      response off
                                         scale?
Reanalyze sample
using second GC
    column.
                             Dilute and reanalyze
                              second aliquot of
                                  sample.
                                       8021A  -  22
                                              Revision  1
                                          September  1994

-------
                                 METHOD 8030A

               ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8030 is used to determine the concentration of the following
volatile organic compounds:
            Compound Name                                   CAS No.a


            Acrolein (Propenal)                             107-02-8
            Acrylonitrile                                   107-13-1


      a  Chemical  Abstract Services Registry Number.

      1.2   Table 1 lists chromatographic conditions and method detection limits
for acrolein and acrylonitrile  in organic-free reagent water.  Table  2 lists the
estimated quantitation limit (EQL) for other matrices.


2.0   SUMMARY OF METHOD

      2.1   Method 8030 provides gas chromatographic conditions for the detection
of the target analytes.  Samples can be analyzed using direct injection or purge-
and-trap  (Method  5030).   Tenax should  be  used as the trap  packing material.
Ground water samples must be analyzed using  Method 5030.  A temperature program
is used in the gas chromatograph to separate the organic compounds.   Detection
is achieved by a flame ionization detector  (FID).

      2.2   The method provides an optional  gas chromatographic column that may
be helpful in resolving the analytes from co-eluting non-target compounds and for
analyte confirmation.
3.0   INTERFERENCES

      3.1   Refer to Methods 5030 and 8000.

      3.2   Samples  can be  contaminated  by  diffusion  of volatile  organics
(particularly chlorofluorocarbons  and  methylene chloride) through  the sample
container  septum  during shipment  and  storage.   A  trip blank  prepared  from
organic-free reagent water and carried through sampling and subsequent storage
and handling can serve as a check on such contamination.
                                   8030A  -  1                         Revision 1
                                                                     July 1992

-------
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1      Gas Chromatograph - Analytical  system complete with gas
      chromatograph suitable for on-column injections or purge-and-trap sample
      introduction and  all  required accessories, including  detectors,  column
      supplies, recorder, gases,  and syringes.  A data  system  for measuring peak
      height and/or peak area is recommended.

            4.1.2      Columns

                 4.1.2.1      Column 1  -  10 ft  x 2  mm ID stainless  steel  or
            glass packed with Porapak-QS  (80/100 mesh) or equivalent.

                 4.1.2.2      Column 2  -  6  ft  x  0.1  in.  ID stainless steel  or
            glass packed with Chromosorb  101 (60/80 mesh) or equivalent.

            4.1.3      Detector - Flame ionization (FID).

      4.2   Sample  introduction  apparatus  -  Refer   to  Method  5030 for  the
appropriate equipment for sample introduction purposes.

      4.3   Syringes - A 5  ml Luer-lok glass hypodermic  and  a 5 mL, gas-tight
with shutoff valve.

      4.4   Volumetric  flasks,  Class  A - Appropriate  sizes  with ground glass
stoppers.

      4.5   Microsyringes   -  10  and   25  p,L  with   a  0.006  in.  ID  needle
(Hamilton 702N, or equivalent) and  a 100  /iL.

      4.6   Analytical balance - 0.0001 g.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall  be used in all tests.  Unless otherwise
indicated, it is intended that all reagents shall  conform to  the  specifications
of the Committee on Analytical Reagents of the  American Chemical  Society, where
such specifications are available. Other grades may be  used, provided it is first
ascertained that  the  reagent is  of sufficiently high purity to permit  its use
without  lessening the accuracy of the  determination.

      5.2   Organic-free reagent water:  All references to  water  in  this method
refer to organic-free reagent water, as defined  in Chapter One.

      5.3   Hydrochloric acid, HC1  - 1:1  (v/v).

      5.4   Sodium  hydroxide, NaOH -   ION  solution.   Dissolve  40 g  NaOH in
organic-free reagent water  and dilute  to  100 ml.

      5.5   Stock standards - Stock solutions may be prepared from pure standard
materials  or purchased  as  certified  solutions.   Prepare  stock standards in

                                   8030A  - 2                        Revision  1
                                                                     July 1992

-------
organic-free  reagent  water using  assayed  liquids.    Because  acrolein  and
acrylonitrile are lachrymators, primary dilutions of these compounds should be
prepared in a hood.

            5.5.1      Place about 9.8  ml  of organic-free reagent water in a 10
      mL tared ground-glass stoppered volumetric  flask.  For acrolein standards
      the water must be adjusted to pH  4-5 using  hydrochloric acid (1:1 v/v) or
      sodium hydroxide  (ION),  if necessary.    Weigh  the flask to  the nearest
      0.0001 g.

            5.5.2      Using a  100 /itL syringe, immediately add two or more drops
      of assayed reference material  to the flask,  then  reweigh.  The liquid must
      fall directly into the water without contacting the neck of the flask.

            5.5.3      Reweigh, dilute  to volume,  stopper,  and  then mix  by
      inverting  the flask several times.    Calculate  the  concentration  in
      milligrams per liter  (mg/L)  from  the net gain  in weight.   When compound
      purity is  assayed  to be  96% or  greater, the weight  may  be  used without
      correction  to  calculate  the  concentration  of  the  stock  standard.
      Commercially prepared stock standards may be used at any concentration, if
      they are certified by the manufacturer or by an independent source.

            5.5.4      Transfer the stock standard solution into a bottle with
      a  Teflon  lined screw-cap.   Store,  with minimal  headspace,  at 4°C and
      protect from light.

            5.5.5      Prepare  fresh standards daily.

      5.6   Secondary dilution  standards - Prepare secondary dilution standards
as needed, in organic-free reagent water, from the stock standard solutions.  The
secondary dilution  standards must  contain the compounds  of  interest,  either
singly or mixed together.  The secondary dilution standards should be prepared
at concentrations such that the  aqueous calibration standards prepared in Section
5.7 will bracket the working range of the analytical system.  Secondary dilution
standards  should be  stored with  minimal headspace,  and  should  be checked
frequently for  signs  of degradation or evaporation, especially just prior to
preparing calibration standards from them.

      5.7   Calibration standards - Prepare calibration standards in organic-free
reagent  water  from  the  secondary  dilution   standards  at a  minimum of five
concentrations.  One of the concentrations  should  be at a concentration near, but
above,  the  method detection  limit.     The   remaining  concentrations  should
correspond to  the  expected range of concentrations found  in  real  samples,  or
should define the working  range  of the GC.   Each standard should  contain each
analyte  for  detection  by this method.   In  order to prepare  accurate aqueous
standard solutions, the following precautions must be observed.

            5.7.1      Use a 25 jitL  Hamilton  702N microsyringe,  or equivalent,
      (variations  in  needle geometry   will  adversely affect  the  ability  to
      deliver reproducible volumes of standards into water).

            5.7.2      Never use pipets  to dilute  or transfer samples or aqueous
      standards.
                                   8030A  -  3                         Revision 1
                                                                     July 1992

-------
            5.7.3      Standards must be prepared daily.

      5.8   Internal standards (if internal standard calibration is used) - To
use this approach, the analyst must select one or more internal standards that
are similar in analytical behavior to  the  compounds  of interest.   The analyst
must further demonstrate that the measurement  of the internal  standard is not
affected by method or  matrix  interferences.   Because of these  limitations, no
internal standard can be suggested that is applicable to all samples.

            5.8.1      Prepare  calibration standards  at  a minimum  of  five
      concentrations for each  parameter of interest,  as described in Section
      5.7.

            5.8.2      Prepare a spiking solution containing each of the  internal
      standards, using the procedures described in Sections  5.5  and 5.6.  It is
      recommended  that  the secondary dilution  standard   be   prepared  at  a
      concentration of 15 mg/L of each internal standard compound.  The  addition
      of 10 /iL of this  standard  to  5.0 ml of  sample  or calibration  standard
      would be equivalent to 30
            5.8.3      Analyze each calibration standard  according  to Section
      7.0, adding 10 nl  of internal  standard spiking solution directly to the
      syringe.

      5.9   Surrogate standards - The analyst should monitor both the performance
of the analytical  system and the effectiveness of the method in dealing with each
sample matrix by spiking each sample, standard, and organic-free reagent water
blank with one or two surrogate compounds  (e.g. compounds similar in analytical
behavior to the analytes  of interest but which are not expected to be present in
the sample) recommended to encompass the range of the temperature program used
in this method.  From stock standard solutions prepared as in Section 5.5, add
a volume to give 750 M9 of  each surrogate to 45 ml  of  organic-free reagent water
contained  in  a 50  mL  volumetric  flask,   mix,  and  dilute   to  volume  for  a
concentration  of 15 ng/juL.    Add  10 juL  of  this surrogate  spiking  solution
directly into the 5 ml syringe with every sample and reference  standard analyzed.
If the internal  standard  calibration  procedure is  used, the surrogate compounds
may be added directly to the internal  standard spiking solution (Section 5.8.2).


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See  the  introductory  material  to  this chapter,  Organic Analytes,
Section 4.1.
7.0   PROCEDURE

      7.1   Volatile compounds are introduced into the gas chromatograph either
by direct injection or heated purge-and-trap (Method 5030).  Method  5030 may be
used directly on ground water samples or low-concentration contaminated soils and
sediments.  For high-concentration soils or sediments, methanolic extraction, as
described in Method 5030, may be  necessary prior to  purge-and-trap  analysis.
                                   8030A - 4                        Revision 1
                                                                     July 1992

-------
      7.2   Gas chromatographlc conditions (Recommended)

            7.2.1      Column 1:

             Helium flow rate =                 30 mL/min
             Temperature program:
                       Initial  temperature =    110°C,  hold for 1.5 minutes
                       Program =                 110°C  to  150°C,   heating  as
                                                rapidly as possible
                       Final  temperature =      150°C,  hold for 20 minutes.

            7.2.2      Column 2:

             Helium flow rate =                 40 mL/min
             Temperature program:
                       Initial  temperature =    80°C, hold for 4 minutes
                       Program =                 80°C to 120°C  at 50°C/"rin
                       Final  temperature =      120°C,  hold for 12 minutes.

      7.3   Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and especially Table 2 for guidance on selecting the  lowest point on
the calibration curve.

            7.3.1      Calibration  must  take  place  using   the  same  sample
      introduction method  that will  be used  to  analyze  actual  samples   (see
      Section 7.4.1).

            7.3.2      The  procedure for internal or external calibration may be
      used.   Refer to Method  8000 for a description of each of  these procedures.

      7.4   Gas chromatographic analysis

            7.4.1      Introduce  volatile compounds into the gas chromatograph
      using  either Method 5030 (heated purge-and-trap method using Tenax as the
      trap packing material) or  the direct  injection method.   If the internal
      standard calibration  technique is  used, add 10 juL  of the  internal standard
      to the sample prior to purging.

                 7.4.1.1      Direct  injection  -  In  very limited applications
            (e.g. aqueous process wastes), direct injection of the sample  into
            the  GC system  with  a  10  pi syringe  may  be appropriate.   The
            detection limit is very high (approximately  10,000 /ug/L); therefore,
            it is only permitted when  concentrations in excess  of 10,000  ng/L
            are expected or for water-soluble  compounds that do  not purge.  The
            system must be  calibrated  by direct injection  (bypassing the purge-
            and-trap device).

            7.4.2      Follow Method  8000  for instructions  on  the  analysis
      sequence,  appropriate  dilutions,  establishing  daily  retention   time
      windows,  and  identification  criteria.     Include   a  mid-concentration
      standard after each group of  10 samples in the analysis sequence.

            7.4.3      Table 1 summarizes the  estimated  retention  times and
      detection limits for  a number of organic compounds analyzable using this

                                  8030A - 5                         Revision 1
                                                                     July  1992

-------
      method.  Figure 1 illustrates the chromatographic separation of acrolein
      and of acrylonitrile using Column 1.

            7.4.4      Record the  sample volume  purged or  injected and  the
      resulting peak sizes (in area units or peak heights).

            7.4.5      Calculation  of concentration  is  covered  in  Method  8000.

            7.4.6      If analytical  interferences  are suspected, or for  the
      purpose  of  confirmation,  analysis   using  the   second   GC  column   is
      recommended.

            7.4.7      If the  response  for  a  peak is  off-scale,   prepare  a
      dilution of the sample with organic-free reagent water.  The dilution must
      be performed on  a  second  aliquot of the sample  which  has  been properly
      sealed and stored prior to use.


8.0  QUALITY CONTROL

      8.1   Refer to  Chapter  One  for specific quality  control  procedures  and
Method 8000 for gas chromatographic procedures.  Quality control  to ensure  the
proper operation of the purge-and-trap device is covered in  Method 5030.

      8.2   Procedures to check the GC system operation are found in Method 8000,
Section 8.6.

            8.2.1      The quality control check sample concentrate  (Method 8000,
      Section 8.6) should contain each parameter  of  interest at a concentration
      of 25 mg/L in water.

            8.2.2      Table  3  indicates the  calibration  and QC  acceptance
      criteria for this method.  Table  4 gives single  laboratory  accuracy  and
      precision  for  the  analytes  of interest.   The contents  of  both  Tables
      should be used to evaluate a  laboratory's ability to perform and generate
      acceptable data by  this method.

      8.3   Calculate  surrogate standard recovery on all  samples, blanks,  and
spikes.  Determine if recovery is within limits (limits established  by performing
QC procedure outlined in  Method 8000, Section 8.10).

            8.3.1      If recovery  is  not  within   limits,   the  following  is
      required.

            •    Check  to be sure  that  there  are  no  errors  in calculations,
                 surrogate  solutions  and   internal  standards.   Also,  check
                 instrument performance.

            •    Recalculate the data and/or reanalyze  the extract if any  of the
                 above checks reveal  a problem.

            •    Reextract and reanalyze the sample  if  none of the above are a
                 problem  or flag the  data as "estimated concentration".


                                   8030A - 6                         Revision 1
                                                                     July 1992

-------
9.0   METHOD PERFORMANCE

      9.1   In  a  single  laboratory,   the  average  recoveries  and  standard
deviations presented in Table 4 were obtained using Method 5030.  Seven replicate
samples were analyzed at each spike concentration.

      9.2   The accuracy and precision obtained will be determined by the sample
matrix, sample introduction technique, and by the calibration procedure used.


10.0  REFERENCES

1.    Bellar, T.A.  and J.J. Lichtenberg,  J. Amer.  Water Works  Assoc., 66(12).
      pp. 739-744,  1974.

2.    Bellar, T.A.  and J.J. Lichtenberg, "Semi-Automated Headspace Analysis of
      Drinking  Waters  and  Industrial  Waters  for Purgeable Volatile Organic
      Compounds," in Van Hall, ed., Measurement of Organic Pollutants in Water
      and Wastewater, ASTM STP 686, pp. 108-129, 1979.

3.    Development and Application of Test  Procedures  for Specific Organic Toxic
      Substances  in Wastewaters,  Category 11:  Purgeables  and Category  12:
      Acrolein,  Acrylonitrile,   and Dichlorodifluoromethane,  Report for  EPA
      Contract 68-03-2635 (in preparation).

4.    Going,  J.,  et  al.,   Environmental  Monitoring Near  Industrial Sites  -
      Acrylonitrile, Office of Toxic Substances, U.S. EPA, Washington, DC, EPA
      560/6-79-003, 1979.

5.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under  the Clean Water Act; Final Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

6.    Kerns, E.H., et al. "Determination of Acrolein  and Acrylonitrile in Water
      by Heated Purge and Trap Technique,"  U.S.  Environmental  Protection Agency,
      Environmental Monitoring and  Support Laboratory,  Cincinnati, Ohio 45268,
      1980.

7.    "Evaluation of Method 603," Final Report for EPA  Contract 68-03-1760 (in
      preparation).
                                   8030A -  7                        Revision 1
                                                                     July 1992

-------
                             TABLE 1.
      CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
                        Retention time (min)             Method detection
Compound                Col. 1        Col.  2              limit8 (/xg/L)
Acrolein
Acrylonitrile
10.6
12.7
8.2
9.8
0.7
0.5
   Based on using purge-and-trap, Method 5030.
                             TABLE 2.
              DETERMINATION OF ESTIMATED QUANTITATION
                LIMITS (EQLs) FOR VARIOUS MATRICES8
         Matrix                              Factor6
         Ground water                             10
         Low-concentration soil                   10
         Water miscible liquid waste             500
         High-concentration soil and sludge     1250
         Non-water miscible waste               1250
         Sample EQLs are highly matrix dependent.   The EQLs  listed  herein
         are provided for guidance and may not always  be achievable.

         EQL =  [Method  detection limit (Table 1)] X  [Factor  (Table  2)].
         For non-aqueous samples, the  factor  is on a  wet-weight  basis.
                             8030A - 8                         Revision 1
                                                                July 1992

-------
                                   TABLE 3.
                    CALIBRATION AND QC ACCEPTANCE CRITERIA3
Analyte
Acrolein
Acrylonitrile
Range
for Q
(M9/L)
45.9 - 54.1
41.2 - 58.8
Limit
for S
(M9/L)
4.6
9.9
Range
for x
(M9/L)
42.9 - 60.1
33.1 - 69.9
Range
P> Ps
(%)
88-118
71-135
Q     =     Concentration measured in QC check sample, in M9/L-
S     =     Standard deviation of four recovery measurements, in /zg/L.
R     =     Average recovery for four recovery measurements, in jig/L.
P, P  =     Percent recovery measured.
      Criteria  from  40  CFR  Part  136  for  Method  603
      assuming a QC check sample concentration of 50
                                        and  were  calculated
                                   TABLE 4.
                   SINGLE LABORATORY ACCURACY AND PRECISION


Parameter
Acrolein





Acrylonitrile





AW
POTW
Spike
cone.
(M9/L)
5.0
50.0
5.0
50.0
5.0
100.0
5.0
50.0
20.0
100.0
10.0
100.0
ASTM Type
Average
recovery
(M9/L)
5.2
51.4
4.0
44.4
0.1
9.3
4.2
51.4
20.1
101.3
9.1
104.0
II water.
Prechlorination secon
Standard
deviation
(M9/L)
0.2
0.7
0.2
0.8
0.1
1.1
0.2
1.5
0.8
1.5
0.8
3.2

dary effluent
Average
percent
recovery
104
103
80
89
2
9
84
103
100
101
91
104

from a munici

Sampl e
matrix
AW
AW
POTW
POTW
IW
IW
AW
AW
POTW
POTW
IW
IW

pal sewage
IW
treatment plant.
Industrial  wastewater  containing  an
reactant.
unidentified  acrolein
                                   8030A -  9
                                                  Revision 1
                                                   July 1992

-------
                   Figure 1
Gas Chromatogram of Acrolein and Acrylonitrile
  Column: Por«pafc.Qs

          110«C lot is mm
          ««tod to 1SO«C
  Ootoctor: FUm«
            <*   «0   78   90   10.5  120   US  ISO
               METCNTlONTlMi. MIN.
                 8030A - 10
Revision 1
 July 1992

-------
                        METHOD  8030A
 ACROLEIN  AND  ACRYLONITRILE BY GAS CHROMATOGRAPHY
      Start
   7  1 Int reduce
compounds into  gas
 chromatograph  by
direct injection or
  purge-and- trap
   (Method 5030)
    7 2 Set gas
   chromatograph
    condition.
   7  3 Calibrate
 (refer to Method
      8000)
  7  4  1 Int r oduce
volatile compounds
    into gas
 chromatograph  by
 purge-and-trap or
 direct injection
742 Folio* Method
 8000 for analysis
  sequence,  etc
7 4  4 Record volume
purged or injected
  and peak sizes
  7  4 5 Calculate
   concent ration
 (refer to Method
      8000)
                        7 4 6 Analyze us ing
                        second GC  column
                        7 4 7 Dilute second
                        aliquot  of sample
                            8030A  -  11
                                          Revision 1
                                           July  1992

-------
                                  METHOD 8031

                      ACRYLONITRUE BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8031 is used to determine  the concentration of acrylonitrile
in water.  This method may also be applicable to other matrices.  The following
compound can be determined by this method:
      Compound Name                                         CAS No.'
      Acrylonitrile                                         107-13-1
      a  Chemical  Abstract Services Registry Number.

      1.2   The estimated quantitation limit of Method 8031 for determining the
concentration of acrylonitrile in water is approximately 10 p.g/1.

      1.3   This method  is  restricted to  use  by or under the  supervision of
analysts  experienced in  the use  of gas  chromatographs and  skilled  in  the
interpretation of gas chromatograms.   Each  analyst must demonstrate the ability
to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   A measured sample volume is micro-extracted with methyl tert-butyl
ether.   The extract  is  separated  by gas  chromatography and measured  with  a
Nitrogen/Phosphorus detector.


3.0   INTERFERENCES

      3.1   Method  interferences  may be caused by contaminants  in  solvents,
reagents, glassware, and  other sample processing hardware that leads to discrete
artifacts and/or elevated baselines in gas chromatograms.  All of these materials
must be routinely demonstrated to be free from interferences under the conditions
of the analysis by running laboratory reagent blanks.

      3.2   Samples can be contaminated by diffusion of volatile organics around
the septum  seal  into the sample during handling and  storage.   A  field blank
should  be  prepared from  organic-free  reagent  water  and carried  through  the
sampling and sample handling protocol to serve as a  check on such contamination.

      3.3   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially  analyzed. To  reduce  carryover,  the


                                   8031  -  1                         Revision  0
                                                                September 1994

-------
sample syringe  must  be rinsed out  between  samples with  solvent.  Whenever an
unusually concentrated  sample is  encountered,  it should  be followed  by  the
analysis of solvent to check for cross contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph system

            4.1.1 Gas  chromatograph,  analytical   system  complete  with  gas
      chromatograph  suitable  for   on-column   injections   and   all   required
      accessories, including detector,  analytical columns, recorder, gases, and
      syringes.  A data system for measuring peak heights and/or peak areas is
      recommended.

            4.1.2 Column:    Porapak Q  -  6 ft., 80/10  Mesh, glass column,  or
      equivalent.

            4.1.3 Nitrogen/Phosphorus detector.

      4.2   Materials

            4.2.1 Grab sample bottles - 40 mL VOA bottles.

            4.2.2 Mixing bottles - 90 ml bottle with a Teflon lined cap.

            4.2.3 Syringes - 10 /uL and 50 juL.

            4.2.4 Volumetric flask (Class A) - 100 ml.

            4.2.5 Graduated cylinder - 50 ml.

            4.2.6 Pipet (Class A) - 5, 15, and 50 ml.

            4.2.7 Vials - 10 ml.

      4.3   Preparation

            4.3.1 Prepare all materials to be used as described in Chapter 4 for
      volatile organics.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests. Unless otherwise
indicated, it is intended that all  reagents  shall conform to  the specifications
of the Committee on Analytical  Reagents of the  American  Chemical Society, where
such specifications are available. Other grades may  be used, provided it  is first
ascertained that  the reagent  is  of sufficiently high  purity to permit its use
without lessening the accuracy of the determination.
                                   8031 - 2                         Revision 0
                                                                September 1994

-------
5.2   General

      5.2.1 Methanol, CH3OH - Pesticide quality,  or equivalent.

      5.2.2 Organic-free reagent water.  All references to water in this
method refer to organic-free reagent water, as defined in Chapter One.

      5.2.3 Methyl  tert-butyl  ether,  CH3Ot-C4H9  -  Pesticide  quality,  or
equivalent.

      5.2.4 Acrylonitrile, H2C:CHCN,  98%.

5.3   Stock standard solution

      5.3.1 Stock standard solutions  - Can  be  prepared from pure standard
materials  or  can be  purchased as certified  solutions.    Prepare  stock
standards  in organic-free reagent water using assayed liquids.

      5.3.2 The stock standard  solution may be  prepared  by volume  or by
weight.   Stock  solutions  must  be replaced  after one year,  or sooner if
comparison with the check standards indicates a problem.

      CAUTION:    Acrylonitrile is toxic.   Standard preparation should be
                  performed  in a laboratory fume hood.

            5.3.2.1     To prepare the stock standard solution by volume:
      inject 10 /uL  of acrylonitrile (98%)  into a 100 mL volumetric flask
      with a syringe.  Make  up to volume with methanol.

            5.3.2.2     To prepare the stock standard solution by weight:
      Place  about 9.8 ml  of  organic-free reagent water  into  a  10  ml
      volumetric  flask before  weighing  the flask and  stopper. Weigh the
      flask and record the weight to  the nearest  0.0001 g.  Add two drops
      of  pure acrylonitrile,  using  a  50 juL  syringe, to  the  flask.   The
      liquid must  fall  directly into the  water, without contacting the
      inside wall  of the  flask.   Stopper the  flask and  then  reweigh.
      Dilute to  volume  with organic-free  reagent  water.   Calculate the
      concentration from the net gain in weight.

5.4   Working standard solutions

      5.4.1 Prepare a minimum of 5 working  standard solutions that cover
the  range of analyte concentrations  expected in  the  samples.   Working
standards of 20,  40, 60, 80,  and 100 M9/L may be  prepared  by injecting 10,
20,  30,  40,  and 50 /uL  of the  stock  standard solution prepared  in Sec.
5.3.2.1 into 5 separate  90 ml mixing bottles containing 40 ml of organic-
free reagent water.

      5.4.2 Inject  15  ml  of methyl   tert-butyl  ether  into  each  mixing
bottle, shake vigorously,  and  let  stand 5  minutes, or  until  layers have
separated.
                             8031 - 3                         Revision 0
                                                          September 1994

-------
            5.4.3 Remove 5 ml of top  layer  by  pipet, and place in a 10 mL vial.

            5.4.4 Keep all standard solutions below 4°C  until  used.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic  Analytes,
Sec. 4.1.


7.0   PROCEDURE

      7.1   Sample Extraction

            7.1.1 Pour 40 mL  of  the sample into a 90 mL mixing bottle.  Pipet 15
      mL of Methyl tert-butyl ether  into the  mixing bottle.   Shake vigorously
      for about 2 min.  and let stand  for about 5 min.  Remove about 5 mL of the
      top layer and store in  a 10 mL  vial.

      7.2   Chromatographic Conditions (Recommended)

      Carrier Gas (He)  flow rate:    35 mL/min.
      Column Temperature:           180° C,  Isothermal
      Injection port temperature:    250° C
      Detector temperature:         250° C
      Detector Current (DC):         18 volts
      Gases:                        Hydrogen,  3 mL/min;   Air, 290 mL/min.

      7.3   Calibration of GC

            7.3.1 On a  daily basis,   inject  3 pi  of  methyl  tert-butyl  ether
      directly into  the  GC to flush  the system.   Also purge the  system with
      methyl tert-butyl  ether injections  between injections of  standards and
      samples.

            7.3.2 Inject  3 jiL of a  sample blank  (organic-free  reagent  water
      carried through the  sample  storage procedures and  extracted  with methyl
      tert-butyl  ether).

            7.3.3 Inject 3 /xL of  at  least  five  standard  solutions:  one should
      be near the detection limit; one should be near,  but below, the expected
      concentrations of the analyte; one should be near, but above, the expected
      concentrations  of  the   analyte.     The   range   of  standard   solution
      concentrations used should not  exceed the working range of the GC system.

            7.3.4 Prepare  a  calibration curve  using  the peak  areas of the
      standards (retention time of acrylonitrile  under  the  conditions of Sec.
      7.2 is  approximately 2.3  minutes).    If the calibration  curve deviates
      significantly from a straight line,  prepare a new calibration curve with
      the existing standards, or, prepare  new standards  and  a new calibration
      curve.    See  Method  8000,  Sec.  7.4.2,  for  additional  guidance  on
      calibration by the external  standard method.


                                   8031 -  4                         Revision 0
                                                                September 1994

-------
      7.4   Sample Analysis

            7.4.1 Inject  3  juL   of  the  sample  extract,   using   the  same
      chromatographic conditions  used to  prepare the standard curve.  Calculate
      the concentration of acrylonitrile in the extract,  using the area of the
      peak, against the calibration curve prepared in Sec.  7.3.4.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and  Method  8000  for  specific quality control
procedures.

      8.2   Prior  to  preparation of  stock  solutions,  methanol  and  methyl
tert-butyl ether reagents should  be analyzed gas chromatographically under the
conditions described in Sec. 7.2, to determine possible interferences with the
acrylonitrile peak.   If the solvent blanks show contamination,  a different batch
of solvents should be used.
9.0   METHOD PERFORMANCE

      9.1   Method 8031 was tested  in a single laboratory over  a period of days.
Duplicate samples and one spiked sample were run for each calculation.  The GC
was calibrated daily.  Results are presented in Table 1.


10.0  REFERENCES

1.    K.L. Anderson,  "The  Determination  of Trace Amounts  of  Acrylonitrile in
      Water by Specific Nitrogen Detector Gas Chromatograph",  American Cynamid
      Report No. WI-88-13, 1988.
                                   8031 - 5                         Revision 0
                                                                September 1994

-------
                        TABLE 1

         SINGLE LABORATORY METHOD PERFORMANCE
                  CONCENTRATION
  SAMPLE          SPIKE (jxg/l.)        % RECOVERY
      A                 60                 100
      B                 60                 105
      C                 40                 86
      D                 40                 100
      E                 40                 88
      F                 60                 94

Average                                    96
                       8031  -  6                          Revision 0
                                                     September 1994

-------
            METHOD 8031
ACRYLONITRILE BY GAS CHROMATOGRAPHY
            C
Start
7.1 .1 Extract 40 mL
of sample with methyl
t-butyl ether in 90 mL
bottle.
1
r
7.2 Set
Chromatographic
conditions.
>
r
7.3.1 Flush GC
system with 30 uL
methyl t-butyl ether.
>
f
7.3.2 Analyze 3 uL
of sample blank.
^
1
7.3.3 - 7.3.4 Establish
calibration curve with
at least 5 stds.
>
r
7.4 Sample Analysis
^
r
                   Stop
              8031 - 7
                               Revision  0
                          September  1994

-------
                                  METHOD 8032

                       ACRYLAMIDE BY GAS CHROMATOGRAPHV
1.0   SCOPE AND APPLICATION

      1.1   Method 8032 is used to determine trace amounts of acrylamide monomer
in aqueous  matrices.   This method  may  be applicable to  other  matrices.   The
following compound can be determined by this method:
      Compound Name                             CAS No.a


      Acrylamide                                79-06-01


      a  Chemical  Abstract Services Registry Number.

      1.2   The method detection limit (MDL) in clean water is 0.032 M9/L-

      1.3   This method  is restricted  to  use  by,  or  under the supervision of,
analysts  experienced  in  the use  of  gas  chromatographs  and  skilled  in  the
interpretation of gas  chromatograms.   Each analyst must demonstrate the ability
to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Method 8032 is based on bromination of the acrylamide double bond.
The reaction product  (2,3-dibromopropionamide)  is  extracted from the reaction
mixture with ethyl acetate,  after salting  out with sodium  sulfate.  The extract
is cleaned up using a Florisil column, and analyzed by gas chromatography with
electron capture detection (GC/ECD).

      2.2   Compound  identification  should  be  supported by   at  least  one
additional qualitative technique.  Analysis using a second gas chromatographic
column  or  gas  chromatography/mass  spectrometry  may  be  used  for  compound
confirmation.


3.0   INTERFERENCES

      3.1   No interference  is observed from sea water or in the presence of 8.0%
of ammonium  ions  derived from  ammonium  bromide.   Impurities  from  potassium
bromide are removed by the Florisil clean up procedure.
                                   8032 - 1                         Revision 0
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatographic System

            4.1.1 Gas chromatograph suitable for on-column injections with all
      required accessories,  including detector,  analytical  columns,  recorder,
      gases, and syringes.  A data  system for measuring peak heights and/or peak
      areas is recommended.

            4.1.2 Column:   2 m  x  3 mm glass column, 5%  FFAP  (free  fatty acid
      polyester) on 60-80 mesh acid washed Chromosorb W,  or equivalent.

            4.1.3 Detector:  electron capture detector.

      4.2   Kuderna-Danish (K-D) apparatus.

            4.2.1 Concentrator tube -  10 ml graduated (Kontes K-570050-1025 or
      equivalent).  A  ground glass stopper  is used  to prevent  evaporation of
      extracts.

            4.2.2 Evaporation  flask  -    500 ml  (Kontes   K-570001-500   or
      equivalent).   Attach  to concentrator tube  with   springs,  clamps,  or
      equivalent.

            4.2.3 Snyder column -   Three  ball macro  (Kontes  K-503000-0121 or
      equivalent).

            4.2.4 Snyder column  -   Two  ball micro (Kontes  K-569001-0219 or
      equivalent).

            4.2.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Separatory funnel - 150 ml.

      4.4   Volumetric flask (Class A)  - 100 ml,  with  ground  glass stopper;
25 ml, amber, with ground glass stopper.

      4.5   Syringe - 5 mL.

      4.6   Microsyringes - 5 jiL,   100 jiL.

      4.7   Pipets (Class A).

      4.8   Glass column (30 cm x  2 cm).

      4.9   Mechanical shaker.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be  used  in  all tests.  Unless otherwise
indicated, it is intended that  all  reagents  shall conform to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where


                                   8032 - 2                          Revision 0
                                                                September  1994

-------
such specifications  are  available.   Other grades may  be  used,  provided it  is
first ascertained that the reagent is of sufficiently high  purity  to  permit  its
use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All  references to  water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Solvents

            5.3.1 Ethyl acetate, C2H5C02C2H5.   Pesticide quality, or equivalent.

            5.3.2 Diethyl ether,  C2H5OC2H5.   Pesticide  quality,  or equivalent.
      Must  be  free  of  peroxides as  indicated by test strips (EM  Quant,   or
      equivalent).   Procedures  for  removal  of peroxides are provided with  the
      test strips.  After cleanup, 20 ml of ethyl alcohol  preservative must  be
      added to each liter of ether.

            5.3.3 Methanol, CH3OH.  Pesticide quality,  or equivalent.

            5.3.4 Benzene, C6H6.  Pesticide  quality,  or equivalent.

            5.3.5 Acetone, CH3COCH3.   Pesticide quality, or equivalent.

      5.4   Saturated bromine water.   Prepare by shaking organic-free reagent
water with bromine and allowing to stand for  1  hour, in the  dark, at 4°C.   Use
the aqueous phase.

      5.5   Sodium sulfate (anhydrous, granular), Na2S04.   Purify  by  heating  at
400°C for 4 hours in a shallow tray, or by precleaning  the sodium sulfate with
methylene chloride.   If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed,  demonstrating that there is no interference from
the sodium sulfate.

      5.6   Sodium thiosulfate, Na2S203,  1 M  aqueous  solution.

      5.7   Potassium bromide,  KBr, prepared  for infrared  analysis.

      5.8   Concentrated hydrobromic acid, HBr, specific gravity  1.48.

      5.9   Acrylamide  monomer,  H2C:CHCONH2,  electrophoresis reagent  grade,
minimum 95% purity.

      5.10  Dimethyl phthalate, C6H4(COOCH3)2,  99.0%  purity.

      5.11  Florisil (60/100 mesh):   Prepare  Florisil by activating at 130°C  for
at least 16 hours.  Alternatively, store Florisil in an oven at 130°C.  Before
use, cool the Florisil  in a desiccator.   Pack 5 g of  the Florisil, suspended  in
benzene, in a glass column (Sec. 4.8).

      5.12  Stock standard solutions

            5.12.1      Prepare a stock standard solution of  acrylamide monomer
      as specified in Sec. 5.12.1.1.   When compound purity  is assayed to be 96%

                                   8032 - 3                          Revision 0
                                                                September 1994

-------
      or greater, the weight  can  be used without correction to  calculate  the
      concentration of the stock standard.  Commercially prepared standards can
      be used at any concentration  if they are certified by the manufacturer or
      by an independent source.

                  5.12.1.1    Dissolve   105.3  mg  of  acrylamide  monomer  in
            organic-free reagent water  in a 100 mL volumetric flask, and dilute
            to the mark with organic-free reagent water. Dilute the solution of
            acrylamide monomer so  as  to obtain  standard  solutions containing
            0.1 - 10 mg/L of acrylamide monomer.

      5.13  Calibration standards

            5.13.1      Dilute the acrylamide stock solution with organic-free
      reagent water to  produce standard solutions  containing 0.1-5  mg/L of
      acrylamide.  Prior to injection the calibration standards are reacted and
      extracted in the same manner as  environmental  samples (Sec. 7).

      5.14  Internal standards

            5.14.1      The suggested  internal  standard is dimethyl phthalate.
      Prepare a  solution  containing 100 mg/L  of dimethyl  phthalate  in  ethyl
      acetate.  The concentration  of dimethyl phthalate in the  sample  extracts
      and calibration standards should  be 4 mg/L.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic  Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Bromination

            7.1.1 Pipet 50 mL  of  sample into a 100 mL  glass  stoppered flask.
      Dissolve 7.5 g of potassium bromide into the sample,  with stirring.

            7.1.2 Adjust the pH of the  solution with  concentrated hydrobromic
      acid until the pH is between 1  and 3.

            7.1.3 Wrap the flask with aluminum foil  in order to exclude light.
      Add 2.5 mL of  saturated bromine water,  with stirring.   Store  the flask and
      contents in the dark, at 0°C, for  at least 1 hour.

            7.1.4 After reacting the solution  for at  least  an hour,  decompose
      the excess of  bromine by adding  1 M sodium thiosulfate  solution, dropwise,
      until the color of the solution is discharged.

            7.1.5 Add 15 g of sodium sulfate, using a magnetic  stirrer to effect
      vigorous stirring.
                                   8032 - 4                         Revision 0
                                                                September 1994

-------
      7.2   Extraction

            7.2.1 Transfer the solution into  a  150 mL separatory funnel.  Rinse
      the reaction flask three times  with  1 ml  aliquots of organic-free reagent
      water.  Transfer the rinsings into the separatory funnel.

            7.2.2 Extract the aqueous solution  with two 10 mL portions of ethyl
      acetate for 2 min each, using a mechanical shaker (240 strokes per min).
      Dry the organic phase with 1 g of sodium sulfate.

            7.2.3 Transfer  the  organic phase  into  a  25  mL  amber  volumetric
      flask.   Rinse the  sodium  sulfate with  three  1.5  ml portions  of ethyl
      acetate and combine the rinsings with the organic phase.

            7.2.4 Add exactly 100 jug  of dimethyl phthalate to the flask and make
      the solution  up to  the 25  ml  mark with  ethyl  acetate.   Inject 5  /xL
      portions of this solution into the gas chromatograph.

      7.3   Florisil cleanup:  Whenever interferences are observed, the samples
should be cleaned up as follows.

            7.3.1 Transfer the dried  extract  into a  Kuderna-Danish  evaporator
      with  15  ml of  benzene.   Evaporate the  solvent  at  70°C under  reduced
      pressure, and concentrate the solution to about 3 mL.

            7.3.2 Add 50  mL of benzene and  subject  the solution to Florisil
      column chromatography at a flow rate of 3 mL/min.  Elute the column first
      with 50 mL of diethyl  ether/benzene  (1:4) at a flow rate of 5 mL/min,  and
      then with  25  mL of  acetone/benzene (2:1) at  a  flow rate of  2  mL/min.
      Discard all of the  first eluate and the initial  9 mL portion of the second
      eluate,  and use  the  remainder  for the determination, using  dimethyl
      phthalate (4 mg/L) as  an internal standard.

            NOTE: Benzene  is  toxic,   and should  be  only be  used  under  a
                  ventilated laboratory hood.

      7.4   Gas chromatographic conditions:

      Nitrogen carrier gas flow rate:     40 mL/min
      Column temperature:                 165°C.
      Injector temperature:               180°C
      Detector temperature:               185°C.
      Injection volume:                    5 ^L

      7.5   Calibration:

            7.5.1 Inject  5  juL of  a  sample blank (organic-free reagent water
      carried through all sample storage,  handling,  bromination and  extraction
      procedures).

            7.5.2 Prepare standard solutions  of acrylamide  as described in Sec.
      5.13.1.   Brominate and extract  each  standard  solution  as described  in
      Sees.  7.1 and 7.2.


                                   8032 -  5                         Revision 0
                                                                September 1994

-------
            7.5.2.1     Inject 5 p,i of each  of  a minimum of five standard
      solutions: one should  be  near the detection limit;  one should be
      near,  but below,  the  expected concentrations of the analyte;  one
      should  be near,   but  above,  the expected  concentrations  of  the
      analyte.

            7.5.2.2     Prepare a calibration curve using the peak areas
      of the standards.  If the calibration curve deviates significantly
      from  a  straight  line,  prepare  a new  calibration  curve  with  the
      existing standards,  or,  prepare  new standards and a new calibration
      curve.   See  Method  8000,  Sec.  7.4.3,  for  additional  guidance on
      calibration by the internal standard method.

            7.5.2.3     Calculate the  response factor  for  each standard
      according to Equation 1.

                  (Ps)  (Mis)
            RF = -              Equation 1
                  (Pis)  (MA)

            RF    =     Response factor
            Ps    =     Peak height of acryl amide
            Mis    =     Amount of internal  standard injected  (ng)
            Pjs    =     Peak height of internal standard
            MA    =     Amount of acryl amide injected (ng)

      7.5.3 Calculate the mean response factor according to Equation 2.
            RF = -                       Equation 2
            RF    =     Mean response factor
            RF    =     Response   factors    from   standard   analyses
                        (calculated in Equation 1)
            n     =     Number of analyses

7.6   Gas chromatographic analysis:

      7.6.1 Inject  5 /iL  portions  of each  sample  (containing 4  mg/L
internal  standard)   into  the  gas  chromatograph.    An  example  GC/ECD
chromatogram is shown in Figure 1.

      7.6.2 The concentration of acrylamide monomer in the sample is given
by Equation 3.

               (PA)   (MJ
      [A] = - — -             Equation 3
             (Pis) (RF) (V()  (V.)
      [A]   =     Concentration of acrylamide monomer in sample  (mg/L)

                             8032 - 6                         Revision 0
                                                          September  1994

-------
            PA    =     Peak height of acryl amide monomer
            Mis    =     Amount of internal standard injected (ng)
            Vs    =     Total volume of sample (mL)
            P^    =     Peak height of internal standard
            RF    =     Mean response factor from Equation 2
            Vi     =     Injection volume
8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and  Method  8000  for specific quality control
procedures.


9.0   METHOD PERFORMANCE

      9.1   The  following  performance  data  have  been  generated  under  the
conditions described in this method:

            9.1.1 The calibration curve  for Method 8032 is linear over the range
      0-5 jug/L of acryl amide monomer.

            9.1.2 The limit of detection for  an aqueous solution is 0.032 M9/L-

            9.1.3 The yields of the  brominated compound are 85.2 +  3.3% and 83.3
      + 0.9%,  at  fortification concentrations of 1.0  and 5.0 M9/U  respectively.

      9.2   Table 1  provides  the recoveries of acrylamide  monomer  from river
water, sewage effluent, and sea water.

      9.3   The recovery of the bromination  product  as a  function of the amount
of potassium  bromide and  hydrobromic acid  added  to the  sample  is  shown  in
Figure 2.

      9.4   The effect of the reaction time on the recovery of the bromination
product is shown  in  Figure 3.  The yield was  constant when the  reaction time was
more than 1 hour.

      9.5   Figure 4 shows the  recovery of the bromination product  as a function
of the initial pH from 1  to  7.35.  The yield was constant within this pH range.
The use of conventional buffer solutions, such as sodium acetate  - acetic acid
solution or phosphate solution, caused a significant decrease in  yield.


10.0  REFERENCES

1.    Hashimoto,   A.,  "Improved  Method  for  the  Determination of  Acrylamide
      Monomer in Water by Means of Gas-Liquid Chromatography with an Electron-
      capture Detector," Analyst, 101:932-938, 1976.
                                   8032 - 7                         Revision 0
                                                                September 1994

-------










co
co
UJ
— i
Q.
•s:
CO
UJ
CH Q
1 1 1 t 1
(_LJ *•— 1
^t ^C
3 Z

y i— *
I-H O Q-
Ct O
i i i i i rv
_ 1 Q-
CQ UJ O
< Q s:
I— >-. o
2: o:
- Q
ry i
o ro
CM
U.
o
2^_
CtL
UJ

o

UJ
oi
































O)
"J3
•1— **-
O
£
ro
14- 0 .,-
it-
CD
o
o
M-
o a
S- E
QJ ro
S_
rO
>

_o

C
0)
> r— E
0 >> 0
U S- C
a> u
ct <

C
0

t— ro S-
i— c a>
rO -i- >
0








S- E O.Q
O) O O
> s- a>
o oo a:





01
3.

BJ
^t
CO
Q
CO
»*
CM

&4—
O

•)->
C
^
o
£j
<




O)
T3
•i— %-
E O
rO ^—
r— O

S- 0

05
^^
13
O
p—
aj
O





CT>
^^
•x^^
"O
CD

•r—
Q.
CO








X
•^
S-
^->
ro
£1

co o en LO o LO
• • • • • •
CO >—< O CM CO CO






*S" CO 00
iii . . .
ill CTt i— i 00
iii en o en
i-H









cvj «d- co oo LO c-»

LO CM CO •— I CO O
00 00 00 00 OO 00








OO LO 1 — . I-H CM ^d"
CO CO t — . CO *3" CM
•— 1 LO ID LO LO LO
o o o o o o








CM rji c\j en en en
CO ^^ fH *^ ^^ ^^
•—I ID 00 ID ID ID
• • • • • •
O O O O O ro







LO O LO O O 
rO O) O) ro
•O S- O> 3 3
C 0) nj r—
ro > S <4- ro
4-> M- a> it- a>
co o; co uj co




































i/i
C
o

+J
ro
{=
•^
E

a>

0)
•o
OJ
"O 0)
•i- +j
E ro
rO O
C -r-
O i—
•i— Q.
Q. CD
O i-

o. at
0 >
E -r-
O >4_
S-
JQ <4_
•i- o
o
i C
CO rO
(1)
CM ^.


a .Q
    9)
 C CT>
 O ~H

 V> S-
 •i- 0)
 > JD
 
    Q.
    0)
    CO
00
CM
CO
O
00

-------
                                   Figure 1
                I
                                                  A


                                                  B
                                    I    10   12   14

                                   Timt/min
                                               16
Typical  gas  chromatograms  of  the bromination  product obtained  from aqueous
acrylamide monomer solution:
   A.
   B.
   BL,
Peaks:

   1.
   I.
   4-7.
Untreated
With Florisil cleanup
Chromatogram of blank,  concentrated five-fold before gas chromatographic
analysis.
    2,3-Dibromopropionamide
    Dimethyl phthalate
    Impurities from potassium bromide
Sample size = 100 ml; acrylamide monomer = 0.1 jug
                                   8032 - 9
                                                            Revision 0
                                                        September  1994

-------
                                  Figure 2
                         0     5    10    IS    20   23

                             Amount of KBr/g ptr SO ml
                         L^_^^^^^BO^^«X>a^^^B^^^^B«^a«^^^U>
                         0     2    «     6    8    10

                            Amount of HBr/ml ptr SO ml
Effect of  (A)  potassium bromide  and  (B) hydrobromic  acid  on the  yield  of
bromination.   Sample  size = 50 ml; acrylamide monomer =  0.25 /ng
                                  8032  -  10
    Revision  0
September 1994

-------
                                   Figure 3
                                                     24
Effect of reaction time on the bromination.  Reaction conditions:

   50 ml of sample;
   0.25 /xg of acrylamide monomer;
   7.5 g of potassium bromide;
   2.5 ml of saturated bromine water

Extraction conditions:

   15 g of sodium sulfate;
   extraction at pH  2;
   solvent = 10 ml of ethyl  acetate (X2)
                                  8032  -  11
    Revision 0
September 1994

-------
                                   Figure 4
                  100
                        0   O    0
                   so

                    012345671

                                       PM
Effect of initial pH on the  bromination.  Reaction and extraction conditions as
in Figure 3.  The pH was adjusted to below 3 with concentrated hydrobromic acid,
and to 4-5 with dilute hydrobromic acid.   Reaction  at pH 6 was  in distilled
water. pH  7.35 was  achieved  by careful addition  of dilute sodium hydroxide
solution.  The  broken line shows the result obtained by the use of sodium acetate
- acetic acid buffer solution.
                                   8032  -  12
    Revision 0
September 1994

-------
                           METHOD  8032
             ACRYLAMIDE  BY GAS  CHROMATOGRAPHY
Start
T
7.1 Bromination
\
7.1 .1 Dissolve 7.5 g KBr into
50 mL sample in flask.
1
7.1 .2 Adjust soln. pH with
concentrated HBr to between
1 and 3.
1
7. 1.3 Wrap soln. flask with
aluminum. Add 2.5 ml satd.
bromine water, stir, store at
0 C for 1 hr
I
7.1 4Add1 M sodium
thiosulfate dropwise to flask to
decompose excess bromine.
1
7.1. 5 Add 15 g sodium
sulfate, and stir



*
7.2 Extraction
1
7.2.1 Transfer flask soln. to
sap. funnel along with rinses.
1
7.2.2 Extract soln. twice w/ethyl
acetate. Dry organic phase
using sodium sulfate.
t
7.2.3 Transfer organic phase
and rinses into amber
glass flask.
1
7.2.4 Add 1 00 ug dimethyl
phthalate to flask, dilute to
mark. Inject 5 uL into GC.
1
7.3 Florisil Cleanup
1
7.3.1 Transfer dried extract to
K-D assembly w/benzene.
Concentrate to 3 ml at 70 C
under reduced pressure.
                                                               I
                                                     7.3.2 Add 50 ml benzene to
                                                     solution. Pass soln. through
                                                     Florisil column. Elutewith
                                                     diethyl ether/benzene, then
                                                     acetone/benzene. Collect
                                                     the second eluton train (less
                                                     initial 9 ml) for analysis.
                            8032  -  13
     Revision  0
September  1994

-------
            METHOD  8032
             continued
       7.4 GC Conditions
        7.5 Calibration
             I
   7.5.1 Inject 5 uL sample blank.
 7.5.2 Brominate and extract std.
 solns. similar to the samples
 .1 Inject 5 uL of each of the
    minimum 5 stds
  .2 Plot peak are vs. [}
  .3 Calculate response factor
    (RF) for each [ ].
   7.5.3 Calculate mean RF from
   eqn.2^	
        7.6 GC Analysis
 7.6.1 Inject 5 uL sample containing
 internal std. into GC.
7.6.2 Calculate acrylamide monomer
concentration in sample using
eqn. 3.
              8032  -   14
      Revision  0
September  1994

-------
                                 METHOD 8040A

                         PHENOLS  BY  GAS  CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8040 is  used to  determine the  concentration  of  various
phenolic compounds.  The following compounds can be determined by this method:
Compound Name
               Appropriate Technique

  CAS No.8   3510   3520    3540  3550  3580
2-sec-Butyl-4,6-dinitrophenol
  (DNBP, Dinoseb)
4-Chloro-3-methylphenol
2-Chlorophenol
Cresols (methyl phenols)
2-Cyclohexyl-4,6-dinitrophenol
2,4-Dichlorophenol
2,6-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
Tetrachlorophenols
Trichlorophenols
2,4,6-Trichlorophenol
   88-85-7   X      ND     ND    ND    X

   59-50-7   X      X      X     X     X
   95-57-8   X      X      X     X     X
 1319-77-3   X      ND     ND    ND    X
  131-89-5   X      ND     ND    ND    LR
  120-83-2   X      X      X     X     X
   87-65-0   X      ND     ND    ND    X
  105-67-9   X      X      X     X     X
   51-28-5   X      X      X     X     X
  534-52-1   X      X      X     X     X
   88-75-5   X      X      X     X     X
  100-02-7   X      X      X     X     X
   87-86-5   X      X      X     X     X
  108-95-2 DC(28)   X      X     X     X
25167-83-3   X      ND     ND    ND    X
25167-82-2   X      X      X     X     X
   88-06-2   X      X      X     X     X
a      Chemical Abstract Services Registry Number.
DC =   Unfavorable distribution coefficient  (number  in  parenthesis is percent
       recovery).
LR =   Low response.
ND =   Not determined.
X  =   Greater than 70 percent recovery by this technique.

       1.2  Table 1 lists the method detection  limit  for the target analytes in
water.  Table 2 lists the estimated quantitation limit (EQL) for all matrices.


2.0    SUMMARY OF METHOD

       2.1  Method 8040 provides gas chromatographic conditions for the detection
of phenolic  compounds.    Prior  to analysis, samples must be  extracted  using
appropriate techniques (see  Chapter  Two  for  guidance).   Both  neat  and diluted
organic  liquids  (Method  3580,  Waste Dilution)  may  be  analyzed by  direct
                                   8040A  -  1
                                 Revision  1
                                  July 1992

-------
injection.  A 2 to 5 /iL  sample  is  injected  into  a gas chromatograph using the
solvent flush technique,  and compounds in the GC effluent are detected by a flame
ionization detector (FID).

       2.2  Method 8040 also  provides  for the preparation of pentafluorobenzyl-
bromide  (PFB)  derivatives,  with  additional  cleanup  procedures for  electron
capture gas  chromatography.   This  is to lower  the  detection limits of some
phenols and to aid the analyst in the elimination of interferences.


3.0    INTERFERENCES

       3.1  Refer to Methods 3500,  3600, and 8000.

       3.2  Solvents, reagents,  glassware, and other sample processing hardware
may yield discrete artifacts and/or elevated baselines causing misinterpretation
of gas chromatograms.  All of these materials must be demonstrated to be free
from interferences, under the conditions of the analysis, by analyzing reagent
blanks.    Specific  selection  of  reagents  and  purification  of   solvents  by
distillation in all-glass systems may be required.

       3.3  Interferences coextracted from samples will vary considerably from
source to  source, depending upon  the  waste  being sampled.   Although general
cleanup techniques are recommended as  part of  this method,  unique  samples may
require additional cleanup.

       3.4  The decomposition of some analytes under basic extraction conditions
has been demonstrated.  Specifically, phenols may react to form tannates.  These
reactions increase with increasing  pH, and are decreased by the  shorter reaction
times available in Method 3510.

       3.5  The flame  ionization  detector (FID)   is very susceptible  to false
positives caused by the presence of hydrocarbons  commonly found in  samples from
waste sites.  The  problem may be minimized by  applying  acid-base cleanup (Method
3650) and/or alumina column chromatography (Method 3611) prior to GC/FID analysis
or  using  the derivatization  technique  and  analyzing by  GC/electron capture
detector.  Initial site investigation should always be  performed utilizing GC/MS
analysis  to  characterize the site and  determine  the  feasibility  of utilizing
Method 8040 with  a GC/FID.


4.0    APPARATUS AND MATERIALS

       4.1  Gas chromatograph

            4.1.1 Gas  Chromatograph  -   Analytical  system  complete  with  gas
       chromatograph  suitable  for  on-column   injections  and  all  required
       accessories, including detectors, column supplies, recorder, gases, and
       syringes.  A data system for measuring  peak areas  and/or peak heights is
       recommended.

            4.1.2 Columns

                  4.1.2.1     Column for underivatized phenols - 1.8 m x  2.0 mm

                                   8040A - 2                         Revision  1
                                                                     July 1992

-------
            ID glass column packed with 1% SP-1240DA on Supelcoport 80/100 mesh,
            or equivalent.

                  4.1.2.2     Column for derivatized  phenols - 1.8 m x 2 mm ID
            glass column  packed  with  5% OV-17 on Chromosorb  W-AW-DMCS 80/100
            mesh, or equivalent.

            4.1.3 Detectors -  Flame ionization (FID) and electron capture (ECD).

       4.2  Reaction vial - 20 ml,  with Teflon lined  screw-cap or crimp top.

       4.3  Volumetric flask,  Class  A  -  Appropriate sizes with  ground-glass
stoppers.

       4.4  Kuderna-Danish (K-D)  apparatus

            4.4.1 Concentrator tube -  10 ml, graduated (Kontes K-570050-1025 or
       equivalent).    Ground-glass  stopper is  used  to prevent  evaporation of
       extracts.

            4.4.2 Evaporation  flask   -  500  ml   (Kontes   K-570001-500   or
       equivalent).    Attach  to  concentrator tube  with  springs,  clamps  or
       equivalent.

            4.4.3 Snyder  column  -  Three  ball  macro  (Kontes  K-503000-0121 or
       equivalent).

            4.4.4 Snyder  column  -  Two ball  micro  (Kontes  K-569001-0219 or
       equivalent).

            4.4.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).

       4.5  Boiling chips - Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

       4.6  Water  bath  - Heated,  with  concentric   ring   cover,  capable  of
temperature control (+ 5°C).   The bath should  be  used in  a  hood.

       4.7  Microsyringe  - 10 juL-

       4.8  Syringe - 5 ml.

       4.9  Balance - analytical, 0.0001 g.


5.0    REAGENTS

       5.1  Reagent grade chemicals shall be used in all tests. Unless otherwise
indicated, it is intended that all  reagents shall conform to the specifications
of the Committee on Analytical Reagents of the American Chemical Society, where
such specifications are available. Other grades may be used,  provided it is first
ascertained that  the  reagent  is  of sufficiently  high  purity to permit its use
without lessening the accuracy of the determination.


                                   8040A - 3                         Revision 1
                                                                     July 1992

-------
       5.2  Organic-free reagent water -  All references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

       5.3  Hexane, CH3(CH2)4CH3 - Pesticide quality or equivalent.

       5.4  2-Propanol, (CH3)2CHOH  - Pesticide  quality or equivalent.

       5.5  Toluene, C6H5CH3 - Pesticide quality or equivalent.

       5.6  Derivatization reagent - Add  1  ml pentafluorobenzyl bromide and 1 g
18-crown-6-ether  to a  50  ml  volumetric  flask  and  dilute  to  volume  with
2-propanol.  Prepare fresh weekly.  This operation  should  be carried out in a
hood.  Store at 4°C and protect from light.

            5.6.1 Pentafluorobenzyl  bromide  (alpha-Bromopentafluorotoluene),
       C6F5CH2Br.  97% minimum purity.

            NOTE:  This chemical is a  lachrymator.

            5.6.2 18-crown-6-ether  (1,4,7,10,13,16-Hexaoxacyclooctadecane)  -
       98% minimum purity.

            NOTE:  This chemical is highly toxic.

       5.7  Potassium carbonate (Powdered), K2C03.

       5.8  Stock standard solutions

            5.8.1 Prepare  stock  standard  solution  at  a  concentration  of
       1000 mg/L  by dissolving  0.0100  g  of  assayed  reference  material  in
       2-propanol and diluting to volume in a  10  ml volumetric flask.  Larger
       volumes can  be  used  at the convenience of the  analyst.   When compound
       purity is assayed  to be 96%  or  greater,  the  weight  can be  used without
       correction  to  calculate  the   concentration  of  the  stock  standard.
       Commercially prepared stock standards can be used at  any concentration if
       they are certified by the manufacturer or by an independent source.

            5.8.2 Transfer the stock standard solutions into bottles with Teflon
       lined screw-caps or  crimp tops.  Store at  4°C  and  protect  from light.
       Stock standards  should be checked  frequently  for signs of degradation or
       evaporation,  especially  just prior to  preparing  calibration standards
       from them.

            5.8.3 Stock standard solutions must be replaced after one year, or
       sooner if comparison with check standards indicates a problem.

       5.9  Calibration standards - Prepare calibration standards at a minimum
of five concentrations  through dilution of  the  stock standards with 2-propanol.
One  of  the concentrations should  be  at  a concentration near,  but  above, the
method detection limit.   The remaining concentrations should correspond to the
expected  range  of concentrations found  in real  samples or  should  define the
working  range  of the  GC.  Calibration  solutions  must be  replaced  after six
months, or sooner,  if comparison with check standards indicates a problem.


                                   8040A  -  4                         Revision 1
                                                                     July  1992

-------
       5.10 Internal standards (if internal standard calibration is used) - To
use this approach, the analyst must select one or more internal standards that
are similar in analytical  behavior to  the  compounds  of interest.   The analyst
must further demonstrate that  the measurement  of the internal  standard is not
affected by method or matrix  interferences.   Because of these limitations, no
internal standard can be suggested that is applicable to all samples.

            5.10.1      Prepare  calibration  standards  at  a  minimum  of  five
       concentrations for each analyte as described in Section 5.9.

            5.10.2      To  each  calibration  standard, add a  known  constant
       amount of one or more  internal  standards,  and dilute to volume with 2-
       propanol.

            5.10.3      Analyze each calibration standard according to Section
       7.0.

       5.11 Surrogate standards - The analyst  should  monitor the performance of
the  extraction,   cleanup   (if  necessary),  and  analytical  system  and  the
effectiveness of the method in dealing with each sample matrix by spiking each
sample, standard, and organic-free reagent  water  blank with  phenolic surrogates
(e.g. 2-fluorophenol  and 2,4,6-tribromophenol)  recommended to encompass the range
of the temperature program used in this method.  Method 3500 details instructions
on the preparation of acid surrogates. Deuterated analogs of analytes should not
be used as  surrogates for gas chromatographic analysis due to  coelution problems.


6.0    SAMPLE COLLECTION, PRESERVATION, AND HANDLING

       6.1  See  the  introductory material  to  this chapter, Organic  Analytes,
Section 4.1.  Extracts must  be stored under refrigeration and analyzed within 40
days of extraction.


7.0    PROCEDURE

       7.1  Extraction

            7.1.1 Refer to Chapter Two for  guidance on  choosing the appropriate
       extraction procedure.  In  general, water samples are extracted  at a pH of
       less than  or equal to 2 with methylene chloride,  using either Method 3510
       or 3520.  Solid samples are extracted using either Method 3540 or 3550,
       and  non-aqueous  samples  using Method  3580.    Extracts obtained  from
       application  of  either  Method 3540  or 3550  should undergo  Acid-Base
       Partition Cleanup, using Method 3650.

            7.1.2 Prior to gas chromatographic analysis, the extraction solvent
       must be exchanged to 2-propanol.  The exchange is performed as follows:

                  7.1.2.1     Following concentration  of the extract  to  1 mL
            using the macro-Snyder column,  allow the apparatus  to cool  and drain
            for at least 10 minutes.
                                   8040A -  5                         Revision 1
                                                                     July 1992

-------
                  7.1.2.2     Remove the micro-Snyder column and rinse its lower
            joint  into  the  concentrator  tube  with  a  minimum  amount of  2-
            propanol.    Adjust  the  extract  volume  to  1.0 ml.   Stopper  the
            concentrator  tube  and  store  refrigerated  at  4°C  if  further
            processing will not be performed immediately.  If the extract will
            be stored longer than two days,  it should be transferred to a vial
            with a Teflon  lined  screw-cap or crimp top.   If the extract requires
            no   further   derivatization   or   cleanup,   proceed   with   gas
            chromatographic analysis.

       7.2  Gas chromatographic conditions (Recommended)

            7.2.1 Column for underivatized phenols -

            Carrier gas (N2)  flow rate:    30 mL/min
            Initial temperature:          80°C
            Temperature program:          80°C to 150°C  at 8°C/min
            Final Temperature:            150°C,  hold until  all compounds have
                                          eluted.

            7.2.2 Column for derivatized phenols -

            Carrier gas (5% methane/95% argon)
            flow rate:                          30 mL/min
            Initial temperature:                200°C
            Temperature program:                isothermal,  hold   until   all
                                                compounds have eluted.

       7.3  Calibration -  Refer to Method 8000 for proper  calibration techniques.
Use Table 1 and especially Table 2 for guidance on selecting  the lowest point on
the calibration curve.

            7.3.1 The procedure for  internal or external  calibration  may be used
       for the underivatized  phenols.  Refer to Method 8000  for  a description of
       each of these procedures.  If derivatization  of the phenols is required,
       the method of external calibration should  be used by injecting five or
       more concentrations of calibration standards that  have also undergone
       derivatization and cleanup prior to instrument calibration.

       7.4  Gas chromatographic analysis

            7.4.1 Refer to Method 8000.   If the internal standard calibration
       technique is used,  add 10 /nL of internal  standard to  the sample prior to
       injection.

            7.4.2 Phenols  are to be determined on a gas chromatograph equipped
       with a flame ionization detector according to the conditions listed for
       the 1%  SP-1240DA column  (Section 7.2.1).   Table  1 summarizes estimated
       retention times and sensitivities that should be  achieved by  this method
       for  clean  water  samples.    Estimated quantitation  limits   for  other
       matrices are list  in Table 2.

            7.4.3 Method  8000 provides  instructions on  the analysis sequence,
       appropriate  dilutions,  establishing  daily retention  time windows, and

                                   8040A - 6                        Revision  1
                                                                     July 1992

-------
identification criteria.  Include a mid-concentration standard after each
group of 10 samples in the analysis sequence.

     7.4.4 An  example  of a GC/FID chromatogram for certain  phenols is
shown in Figure  1.   Other packed  or  capillary (open-tubular)  columns,
chromatographic conditions, or detectors may be used if the requirements
of Section 8.2 are met.

     7.4.5 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

     7.4.6 Using either  the internal  or  external  calibration procedure
(Method 8000), determine  the identity and quantity  of each component peak
in the sample  chromatogram which  corresponds  to the compounds  used for
calibration purposes.  See Method 8000 for calculation equations.

     7.4.7 If  peak detection  using the SP-1240DA  column  with the flame
ionization detector is  prevented by interferences,  PFB derivatives of the
phenols  should be  analyzed  on  a gas chromatograph equipped with  an
electron capture detector according to the conditions listed for the 5%
OV-17 column (Section 7.2.2).   The derivatization and cleanup procedure
is outlined in Sections  7.5 through 7.6.   Table 3 summarizes estimated
retention times for derivatives of some phenols using the conditions of
this method.

     7.4.8 Figure 2  shows  a GC/ECO chromatogram  of PFB  derivatives of
certain phenols.

     7.4.9 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

     7.4.10      Determine the  identity and quantity of  each component
peak in the sample chromatogram which corresponds  to the compounds used
for calibration purposes.  The method of external  calibration should be
used (see Method 8000 for guidance). The concentration of the individual
compounds in the sample  is calculated  as  follows:
     Concentration
where:
     A     =     Mass of underivatized phenol represented by area of peak
                 in  sample chromatogram,  determined from calibration
                 curve (see Method 8000), ng.

     Vt    =     Total amount of column eluate or combined fractions from
                 which V, was taken,  juL.

     B     =     Total volume of hexane added in Section 7.5.5, ml.

     D     =     Total   volume   of    2-propanol   extract  prior   to
                 derivatization, ml.

                            8040A  - 7                         Revision 1
                                                              July 1992

-------
            V,     =     Volume injected, nl.

            X     =     Volume of water extracted,  mL,  or weight of nonaqueous
                        sample extracted, g, from Section 7.1.   Either the dry
                        or wet weight  of the  nonaqueous  sample may  be  used,
                        depending upon the specific application of the data.

            C     *     Volume of hexane sample solution added to cleanup column
                        (Method 3630),  mL.

            E     =     Volume   of   2-propanol    extract   carried   through
                        derivatization in Section 7.5.1, ml.

       7.5  Derivatization - If interferences prevent measurement of peak area
during analysis  of the  extract  by flame  ionization gas chromatography,  the
phenols must be derivatized and analyzed by electron capture gas chromatography.

            7.5.1 Pipet  a 1.0 ml aliquot  of the  2-propanol  stock  standard
       solution  or of the  sample extract into a glass  reaction vial.  Add 1.0 ml
       derivatization  reagent (Section 5.3).    This   amount  of  reagent  is
       sufficient to derivatize a solution whose total phenolic content does not
       exceed 300 mg/L.

            7.5.2 Add  approximately  0.003  g  of  potassium  carbonate to  the
       solution  and shake gently.

            7.5.3 Cap the mixture and heat  it  for 4 hours at 80°C  in a hot water
       bath.

            7.5.4 Remove the solution from  the hot water  bath  and  allow it to
       cool.

            7.5.5 Add 10 ml  hexane to the reaction vial  and shake vigorously for
       1  minute.   Add  3.0 mL organic-free reagent  water  to  the reaction vial and
       shake for 2 minutes.

            7.5.6 Decant the organic layer into a  concentrator tube and cap with
       a  glass stopper.  Proceed  with  cleanup procedure.

       7.6  Cleanup

            7.6.1 Cleanup of the derivatized extracts takes place using Method
       3630 (Silica Gel Cleanup),  in which  specific instructions  for cleanup of
       the derivatized phenols appear.

            7.6.2 Following column cleanup,  analyze the samples  using GC/ECD, as
       described starting in Section 7.4.7.


8.0    QUALITY CONTROL

       8.1  Refer  to  Chapter  One  for   specific  quality control  procedures.
Quality control  to validate sample extraction is  covered in Method 3500 and in


                                   8040A -  8                        Revision  1
                                                                     July 1992

-------
the extraction method used.   If extract cleanup was performed, follow the QC in
Method 3600 and in the specific cleanup method.

       8.2  Procedures to check the GC system operation are found  in Method 8000,
Section 8.6.

            8.2.1 The quality  control  check sample concentrate  (Method 8000,
       Section 8.6) should contain each analyte  of interest at a concentration
       of 100 mg/L in 2-propanol.

            8.2.2 Table 4 indicates the calibration and QC acceptance criteria
       for  this  method.    Table  5  gives  method  accuracy  and precision  as
       functions of concentration  for the analytes. The contents of both tables
       should be used to evaluate a laboratory's  ability to perform and generate
       acceptable data by this method.

       8.3  Calculate surrogate standard  recovery  on  all  samples,  blanks,  and
spikes.   Determine  if the  recovery is within  limits  (limits  established  by
performing QC procedures outlined in Method 8000, Section 8.10).

            8.3.1 If recovery  is not within limits, the following is required.

            •     Check to  be  sure  that  there are no  errors  in calculations,
                  surrogate  solutions  and  internal  standards.   Also,  check
                  instrument performance.

            •     Recalculate  the data and/or reanalyze the  extract  if any of
                  the above checks reveal  a problem.

            •     Reextract and reanalyze  the  sample  if none of the above are a
                  problem or flag the data as "estimated concentration."


9.0    METHOD PERFORMANCE

       9.1  The method was tested  by 20 laboratories using organic-free reagent
water, drinking water,  surface  water, and three industrial wastewaters spiked at
six concentrations over the  range 12 to 450 /zg/L.   Single operator precision,
overall precision, and method accuracy were  found to be  directly related to the
concentration of the analyte and essentially independent of the sample matrix.
Linear equations to describe  these relationships  for a flame ionization detector
are presented in Table 5.

       9.2  The accuracy and precision obtained  will  be affected by the sample
matrix, sample-preparation technique, and calibration procedures used.


10.0   REFERENCES

1.     Development and Application of Test Procedures  for Specific Organic Toxic
       Substances in  Wastewaters.   Category 3 -  Chlorinated  Hydrocarbons  and
       Category  8  -  Phenols.    Report  for  EPA  Contract  68-03-2625  (in
       preparation).


                                  8040A - 9                        Revision 1
                                                                     July 1992

-------
U.S. EPA 40 CFR  Part  136,  "Guidelines Establishing Test Procedures for
the Analysis  of  Pollutants Under  the Clean Water Act;  Final  Rule and
Interim Final  Rule and Proposed Rule," October 26, 1984.

"Determination of Phenols  in  Industrial  and  Municipal  Wastewaters,"
Report for EPA Contract 68-03-2625 (in preparation).

"EPA Method Validation Study Test Method 604 (Phenols)," Report for EPA
Contract 68-03-2625 (in preparation).

Kawahara,  F.K.  "Microdetermination  of  Derivatives  of  Phenols  and
Mercaptans by Means of Electron Capture Gas Chromatography," Analytical
Chemistry, 40, 1009, 1968.

Burke, J.A.  "Gas Chromatography  for Pesticide  Residue  Analysis; Some
Practical Aspects," Journal  of the  Association  of Official  Analytical
Chemists, 48,  1037, 1965.
                            8040A -  10                        Revision 1
                                                               July 1992

-------
                                   TABLE 1.
                FLAME IONIZATION GAS CHROMATOGRAPHY OF PHENOLS8
                                                                 Method
                                             Retention  time       Detection
Analyte                                      (minutes)            limit


2-sec-Butyl-4,6-dinitrophenol (DNBP)
4-Chloro-3-methylphenol                       7.50                   0.36
2-Chlorophenol                                1.70                   0.31
Cresols (methyl phenols)
2-Cyclohexyl-4,6-dinitrophenol
2,4-Dichlorophenol                            4.30                   0.39
2,6-Dichlorophenol
2,4-Dimethylphenol                            4.03                   0.32
2,4-Dinitrophenol                            10.00                  13.0
2-Methyl-4,6-dinitrophenol                   10.24                  16.0
2-Nitrophenol                                 2.00                   0.45
4-Nitrophenol                                24.25                   2.8
Pentachlorophenol                            12.42                   7.4
Phenol                                        3.01                   0.14
Tetrachlorophenols
Trichlorophenols
2,4,6-Trichlorophenol                         6.05                   0.64


8 -  1% SP-1240DA on Supelcoport 80/100 mesh column.
                                   TABLE 2.
                    DETERMINATION OF  ESTIMATED QUANTITATION
                      LIMITS (EQLs) FOR VARIOUS MATRICES8
   Matrix                                                   Factor6


Ground water                                                     10
Low-concentration soil by sonication with GPC cleanup           670
High-concentration soil and sludges by sonication            10,000
Non-water miscible waste                                    100,000


a  Sample EQLs are highly matrix-dependent.  The EQLs listed herein  are provided
  'for guidance and may not always be achievable.

b  EQL =  [Method  detection limit  (Table  1)]  X [Factor  (Table  2)].   For non-
   aqueous samples,  the factor is on a wet-weight basis.
                                  8040A - 11                        Revision 1
                                                                     July 1992

-------
                                 TABLE 3.
          ELECTRON  CAPTURE  GAS  CHROMATOGRAPHY  OF  PFB  DERIVATIVES8


Parent compound
4-Ch1oro-2-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl -4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Retention
time
(min)
4.8
3.3
5.8
2.9
46.9
36.6
9.1
14.0
28.8
1.8
7.0
Method
detection
limit (M9/L)
1.8
0.58
0.68
0.63


0.77
0.70
0.59
2.2
0.58
- 5% OV-17 on Chromosorb W-AW-DMCS 80/100 mesh column.
                                8040A  -  12                        Revision  1
                                                                   July  1992

-------
                               TABLE 4.
                        QC ACCEPTANCE CRITERIA8


Analyte
4-Chloro-3-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(M9/L)
16.6
27.0
25.1
33.3
25.0
36.0
22.5
19.0
32.4
14.1
16.6
Range
for x
(M9/L)
56.7-113.4
54.1-110.2
59.7-103.3
50.4-100.0
42.4-123.6
31.7-125.1
56.6-103.8
22.7-100.0
56.7-113.5
32.4-100.0
60.8-110.4
Recovery
Range
(%)
99-122
38-126
44-119
24-118
30-136
12-145
43-117
13-110
36-134
23-108
53-119
s

x

a
= Standard deviation of four recovery measurements, in

= Average recovery for four recovery measurements, in jug/L.

Criteria from  40  CFR Part 136  for  Method 604.   These  criteria are based
directly upon the method performance data in Table 5.  Where necessary, the
limits  for  recovery have  been  broadened  to  assure applicability  of the
limits to concentrations below those used to develop Table 5.
                              8040A - 13
                                                                    Revision 1
                                                                     July 1992

-------
                                   TABLE 5.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION8
Analyte
4-Chloro-3-methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2, 4-Dimethyl phenol
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Accuracy, as
recovery, x'
(M9/L)
0.87C-1.97
0.83C-0.84
0.81C+0.48
0.62C-1.64
0.84C-1.01
0.80C-1.58
0.81C-0.76
0.46C+0.18
0.83C+2.07
0.43C+0.11
0.86C-0.40
Single analyst
precision, s '
(M9/D
O.llx-0.21
O.lSx+0.20
0.17X-0.02
O.SOx-0.89
0.15X+1.25
0.27X-1.15
O.lBx+0.44
0.17X+2.43
0.22X-0.58
0.20X-0.88
O.lOx+0.53
Overal 1
precision,
S' (ng/D
0.16X+1.41
0.21X+0.75
O.lSx+0.62
0.25x+0.48
0.19X+5.85
0.29X+4.51
0.14x+3.84
0.19X+4.79
0.23X+0.57
O.Ux+0.77
0.13x+2.40
X'
S'


C

x
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C,  in /xg/L.

Expected single  analyst  standard deviation of measurements  at an
average concentration of x, in M9/L-

Expected interlaboratory standard deviation of measurements  at an
average concentration found of x, in M9/L*

True value for the concentration, in /xg/L.

Average recovery found for  measurements  of samples  containing a
concentration of C, in M9/L.
8From 40 CFR Part 136 for Method 604.
                                  8040A  -  14
                                                        Revision  1
                                                         July  1992

-------
          Figure 1
Gas  Chromatogram  of  Phenols
   Column: 1% SM2400A on Soptieooort
   Program: 80°C 0 Minute* 8°/Minuta to 1SO°C
   Dttactor: Flama lonisation
8      12      16      20
 RETENTION TIME (MINUTES)
24
21
          8040A  -  15
                Revision  1
                 July 1992

-------
                Figure 2
Gas Chromatogram of PFB Derivatives of Phenols
             Column: 5% OV-17 on OtremoMrb W-AW
                      200°C
                    Electron Cteturt
                                       A_
            12    16    20     24
         RETENTION TIME (MINUTES)
21     32
                8040A  - 16
        Revision 1
         July 1992

-------
                       METHOD 8040A
             PHENOLS BY  GAS  CHROMATOGRAPHY
 7  1 1 Choose
 appropriate
 ex t raction
 method (refer
 to Chapter 2)
    7 1 2
  Exchange
  OKt raction
  solvent to
  2 -propanol
  7  2 Set gas
chromatography
  conditions
 7 3 Refer to
  Method 8000
  for proper
  cal ibra to. on
  techniques
3 1 la N.
atization x. Yea
phenols j— — — +
quired' /

731 Inject at
least 5
concentrations
of ca 1 ibra tion
s tandar ds


 7 4 Perform
 CC analysis
 (see Method
   8000)
 7  1 analyze
using GC/FID
                       8040A  - 17
                                        Revision  1
                                         July  1992

-------
          METHOD 8040A
           (Continued)
                           7 5  Prepare
                           derivatives
 749 Record
 sample volume
 injected and
  peak sizes
 7  k Cleanup
using Method
    3630
    7 4 10
Identitify and
quantitate each
component peak
7 4  7 Analyze
    PFB
 deriva tives
using CC/ECD
    7 4 10
   Calculate
 concentration
     Stop
            8040A  - 18
                             Revision 1
                              July 1992

-------
                                 METHOD 8060

                              PHTHALATE ESTERS
1.0  SCOPE AND APPLICATION

     1.1  Method 8060  Is  used  to  determine  the  concentration  of various
phthalate esters.  Table 1 indicates  compounds that may be determined by this
method and lists  the  method  detection  limit  for  each compound in reagent
water.   Table  2  lists  the  practical  quantisation  limit  (PQL) for other
matrices.
2.0  SUMMARY OF METHOD

     2.1  Method  8060  provides   gas   chromatographic  conditions  for  the
detection of ppb levels of  phthalate  esters.    Prior to use of this method,
appropriate sample extraction techniques must be  used.  Both neat and diluted
organic liquids  (Method  3580,  Waste  Dilution)  may  be  analyzed by direct
injection.  A 2-  to  5-uL  aliquot  of  the  extract  is  injected into a gas
chromatograph (GC) using the solvent flush  technique, and compounds in the GC
effluent are  detected  by  an  electron  capture  detector  (ECD)  or a flame
ionization detector  (FID).  Ground water samples should be determined by ECD.

     2.2  The method provides a second  gas chromatographic column that may be
helpful in resolving the analytes  from  interferences  that may occur and for
analyte confirmation.


3.0  INTERFERENCES

     3.1  Refer to Methods  3500, 3600, and 8000.

     3.2  Phthalate  esters  contaminate many   types  of products commonly found
in the  laboratory.   The  analyst  must  demonstrate that no phthalate  residues
contaminate the sample or   solvent   extract   under the conditions  of analysis.
Plastics, in particular, must  be  avoided because phthalates are  commonly  used
as plasticizers and  are  easily  extracted   from  plastic materials.  Serious
phthalate contamination may result at  any  time if consistent quality control
is not  practiced.

     3.3  Solvents,  reagents,  glassware, and  other sample processing  hardware
may   yield   discrete    artifacts    and/or   elevated   baselines   causing
misinterpretation  of  gas   chromatograms.      All  these  materials   must  be
demonstrated to   be  free   from  interferences,  under  the  conditions of the
analysis, by analyzing  method blanks.    Specific   selection of reagents and
purification of  solvents by distillation in all-glass systems may be  required.

     3.4   Interferences coextracted  from  samples  will vary considerably  from
source  to source,  depending upon  the  waste  being sampled.  Although general
cleanup techniques are recommended as  part  of  this method, unique samples may
require additional cleanup.


                                  8060 - 1
                                                         Revision      0
                                                         Date  September 1986

-------
TABLE 1.  RETENTION TIME AND DETECTION LIMIT INFORMATION FOR PHTHALATE ESTERS

                                Retention time (mln)      Method detection
                                                            limit (ug/L)
Compound                         Col. la    Col. 2b        ECD        FID
Benzyl butyl phthalate
B1 s (2-ethyl hexyl ) phthal ate
Di-n-butyl phthalate
D1 ethyl phthalate
Dimethyl phthalate
Di-n-octyl phthalate
*6.94
*8.92
8.65
2.82
2.03
*16.2
**5.11
**10.5
3.50
1.27
0.95
**8.0
0.34
2.0
0.36
0.49
0.29
3.0
15
20
14
31
19
31
     aColumn 1:  Supelcoport 100/120  mesh  coated with 1.5% SP-2250/1.95% SP-
     2401 packed in a 180-cm x 4-mm  I.D.  glass column with carrier gas at 60
     mL/min flow rate.  Column temperature  is 180*C, except where * indicates
     220*C.  Under these conditions the  retention  time of Aldrin is 5.49 min
     at 180*C and 1.84 min at 220*C.

     bColumn 2:  Supelcoport 100/120 mesh with 3% OV-1 in a 180-cm x 4-mm I.D.
     glass column with carrier gas at 60 mL/min flow rate.  Column temperature
     is 200*C, except where **  indicates  220*C.   Under these conditions the
     retention time of Aldrin is 3.18 min at 200*C and 1.46 min at 220*C.
 TABLE  2.   DETERMINATION  OF PRACTICAL  QUANTITATION  LIMITS  (PQL)  FOR  VARIOUS
           MATRICESa


     Matrix                                                    Factorb


 Ground water                                                     10
 Low-level  soil  by sonication with GPC cleanup                   670
 High-level soil and sludges by sonication                    10,000
 Non-water miscible waste                                    100,000


      aSample PQLs are highly  matrix-dependent.     The  PQLs listed herein are
      provided for guidance and may not always be achievable.

      bPQL = [Method detection limit (Table 1)] X [Factor (Table 2)].  For non-
      aqueous samples, the factor is on a wet-weight basis.
                                   8060 - 2
                                                          Revision
                                                          Date  September 1986

-------
4.0  APPARATUS AND MATERIALS
     4.1  Gas chromatograph;
          4.1.1  Gas  chromatograph:    Analytical  system  complete  with gas
     chromatograph  suitable  for   on-column   injections  and  all  required
     accessories, including detectors,  column  supplies, recorder, gases, and
     syringes.  A data system for  measuring peak areas and/or peak heights is
     recommended.
          4.1.2  Columns:
               4.1.2.1  Column 1:  1.8-m x  4-mm  I.D. glass column packed with
          1.5%  SP-2250/1.95%   SP-2401   on   Supelcoport   100/120  mesh  or
          equivalent.
               4.1.2.2  Column 2:  1.8-m x  4-mm  I.D. glass column packed with
          3% OV-1 on Supelcoport 100/120 mesh or  equivalent.
          4.1.3  Detectors:  Flame ionization (FID) or electron capture (ECD).
     4.2  Volumetric flask;  10-, 50-, and 100-mL, ground-glass stopper.
     4.3  Kuderna-Danish  (K-D) apparatus:
          4.3.1  Concentrator tube:   10-mL, graduated  (Kontes K-570050-1025 or
     equivalent).  Ground-glass  stopper  is  used  to  prevent evaporation of
     extracts
          4.3.2  Evaporation   flask:      500-mL  (Kontes   K-570001-500  or
     equivalent).  Attach  to concentrator tube with springs.
          4.3.3  Snyder column:    Three-ball  macro   (Kontes K-503000-0121 or
     equivalent).
          4.3.4  Snyder   column:    Two-ball  micro   (Kontes  K-569001-0219 or
     equivalent).
     4.4  Boiling chips;   Solvent extracted,  approximately  10/40 mesh  (silicon
 carbide or  equivalent).
     4.5  Water  bath;     Heated,  with  concentric   ring   cover,   capable  of
 temperature control  (+5*C).  The bath should  be  used  in  a hood.
     4.6  Mlcrosyringe;   10-uL.
     4.7  Syringe;   5-mL.
     4.8  Vials;  Glass,  2- and  20-mL capacity with Teflon-lined  screw cap.
                                   8060 -  3
                                                          Revision
                                                          Date   September  1986

-------
5.0  REAGENTS

     5.1  Solvents:    Hexane,   acetone,  isooctane  (2,2,4-trimethylpentane)
(pesticide quality or equivalent).

     5.2  Stock standard solutions;

          5.2.1  Prepare stock standard solutions  at  a concentration of 1.00
     ug/uL by dissolving 0.0100 g  of  assayed reference material in isooctane
     and diluting to volume in a  10-mL  volumetric flask.  Larger volumes can
     be used at the  convenience  of  the  analyst.    When compound purity is
     assayed to be 96% or greater,  the  weight can be used without correction
     to calculate  the  concentration  of  the  stock  standard.  Commercially
     prepared stock standards can  be  used  at  any concentration if they are
     certified by the manufacturer or by an independent source.

          5.2.2  Transfer  the  stock  standard  solutions  into Teflon-sealed
     screw-cap bottles.  Store at 4*C and protect from light.  Stock standards
     should be checked  frequently  for  signs  of degradation or evaporation,
     especially just prior to preparing  calibration standards from them.

          5.2.3  Stock standard solutions must be  replaced after one year, or
     sooner  if comparison with check standards indicates  a problem.

     5.3  Calibration standards:  Calibration  standards  at a minimum of five
concentrationlevelsshouldbe prepared  through  dilution   of  the  stock
standards with isooctane.   One   of  the concentration   levels  should be at a
concentration near, but  above,   the  method  detection   limit.  The remaining
concentration levels should correspond to the expected range of  concentrations
found   in   real   samples  or  should  define  the  working  range  of  the GC.
Calibration   solutions  must  be   replaced  after  six  months,  or  sooner if
comparison  with  check standards indicates a problem.

     5.4   Internal  standards  (if  internal  standard  calibration is used);  To
use this  approach,  the analyst must select one or more internal  standards that
are similar in analytical behavior to  the compounds of interest.  The analyst
must further demonstrate that the measurement  of  the internal  standard  is not
affected  by method or matrix  interferences.   Because of  these  limitations, no
 internal  standard  can be suggested that  is applicable to  all samples.

           5.4.1   Prepare  calibration    standards    at    a   minimum  of five
     concentration  levels   for   each  analyte   of   interest   as described in
      Paragraph  5.3.

           5.4.2   To each  calibration  standard,  add   a known  constant  amount of
     one  or more internal  standards,  and dilute  to  volume with  isooctane.

           5.4.3   Analyze each calibration standard according  to Section  7.0.

      5.5   Surrogate standards;   The  analyst   should monitor the performance of
 the extraction,  cleanup  (when   used),   and   analytical   system and the  effec-
 tiveness  of the method  in  dealing   with  each  sample matrix by  spiking each


                                   8060 - 4
                                                          Revision       0
                                                          Date  September 1986

-------
sample, standard, and reagent water  blank  with  one or two surrogates (e.g.,
phthalates that are not expected to be In the sample) recommended to encompass
the range of  the  temperature  program  used  in  this  method.   Method 3500,
Section 5.3.1.1,  details  instructions  on  the  preparation  of base/neutral
surrogates.  Deuterated analogs of  analytes  should not be used  as surrogates
for gas chromatographic analysis due to coelution problems.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See the introductory  material  to  this  chapter, Organic Analytes,
Section 4.1.  Extracts must be  stored under refrigeration and analyzed within
40 days of extraction.


7.0  PROCEDURE

     7.1   Extraction:

           7.1.1  Refer to Chapter Two  for guidance on choosing the appropriate
     extraction  procedure.    In  general,  water  samples  are  extracted at a
     neutral, or as  is, pH  with  methylene chloride, using either Method 3510
     or  3520.   Solid samples  are extracted using either Method 3540 or 3550.

           7.1.2 Prior to gas chromatographic analysis, the extraction solvent
     must  be  exchanged to hexane.    The  exchange is performed during the K-D
     procedures listed in all   of   the extraction   methods.   The exchange is
     performed  as  follows.

                7.1.2.1   Following  K-D  of  the methylene  chloride  extract  to
           1 mL  using the macro-Snyder  column,   allow the  apparatus to  cool and
           drain for  at least  10 min.

                7.1.2.2  Momentarily remove the   Snyder  column,   add   50 mL of
           hexane,  a  new boiling chip,  and  reattach the  macro-Snyder column.
           Concentrate the extract  using  1  mL   of  hexane  to  prewet the  Snyder
           column.  Place the   K-D   apparatus  on  the   water  bath so  that the
           concentrator tube  is partially   immersed   in  the  hot  water.   Adjust
           the vertical position of the apparatus and the  water temperature, as
           required,  to complete concentration in 5-10 min.  At the proper rate
           of  distillation the balls of the  column  will  actively chatter, but
           the chambers will  not flood.    When the apparent volume of  liquid
           reaches  1  mL,  remove the   K-D   apparatus   and   allow it to drain and
           cool  for at least  10 min.    The extract will be handled differently
           at  this  point, depending  on  whether   or   not cleanup  is needed.  If
           cleanup  is not required,  proceed   to   Paragraph  7.1.2.3.  If cleanup
           is  needed, proceed  to Paragraph  7.1.2.4.

                7.1.2.3   If cleanup  of  the  extract   is  not  required, remove the
           Snyder column  and   rinse   the   flask   and   its   lower  joint  into the
           concentrator tube   with   1-2 mL   of   hexane.     A  5-mL syringe is
           recommended for this operation.  Adjust the extract volume to


                                   8060 -  5
                                                          Revision     0
                                                          Date September 1986

-------
         10.0 ml.  Stopper  the  concentrator  tube and store refrigerated at
         4*C if further processing will not be performed immediately.  If the
         extract  will  be  stored  longer   than  two  days,  it  should  be
         transferred to a  Teflon-sealed  screw-cap  vial.   Proceed with gas
         chromatographic analysis.

              7.1.2.4  If cleanup  of  the  extract  is  required, remove the
         Snyder column and  rinse  the  flask  and  its  lower joint into the
         concentrator tube with a minimum  amount  of hexane.  A 5-mL syringe
         is recommended for this operation.   Add a clean boiling chip to the
         concentrator tube and attach a two-ball micro-Snyder column.  Prewet
         the column by adding about 0.5 mL  of  hexane to the top.  Place the
         micro-K-D  apparatus  on  the   water   bath    (80*C)  so  that  the
         concentrator tube is partially  immersed  1n  the hot water.  Adjust
         the vertical position of the apparatus and the water temperature, as
         required,  to complete concentration  in 5-10 min.  At the proper rate
         of distillation  the balls of  the  column will  actively chatter, but
         the chambers will not  flood.    When  the apparent volume of liquid
         reaches  0.5 ml,  remove the K-D  apparatus  and  allow it to drain and
         cool  for at  least  10 min.

               7.1.2.5   Remove the micro-Snyder  column and  rinse the flask and
          its  lower joint  into the  concentrator  tube  with  0.2 ml of hexane.
         Adjust the extract  volume to  2.0  ml  and proceed with either Method
          3610  or 3620.

     7.2  Gas   chromatpgraphy  conditions   (Recommended);    The   analysis for
phthalate esters may  Be  conducted   using   either  a   flame   ionization  or  an
electron capture detector.  The  ECD may,  however,  provide substantially  better
sensitivity.

          7.2.1  Column 1:  Set 5%  methane/95%  argon  carrier gas  flow at  60
     mL/min flow rate.  Set column temperature at 180*C isothermal.

          7.2.2  Column 2:  Set 5%  methane/95%  argon  carrier gas  flow at  60
     mL/min flow rate.  Set column temperature at 200*C isothermal.

     7.3  Calibration;    Refer   to   Method   8000  for  proper  calibration
techniques"!  Use Table 1 and especially  Table 2 for guidance on selecting the
lowest point on the calibration curve.
     7.3.1  The procedure for  internal   or  external
used.    Refer  to  Method  8000  for  a  description
procedures.
                                                            calibration may be
                                                            of  each  of these
          7.3.2  If cleanup is performed  on  the  samples, the analyst should
     process a series  of  standards  through  the  cleanup procedure and then
     analyze the samples by GC,    This  will confirm elution patterns and the
     absence of interferents from the reagents.
                                  8060 - 6
                                                         Revision      0
                                                         Date  September 1986

-------
    7.4  Gas chromatographic analysis;

         7.4.1   Refer to Method 8000.     If the Internal standard calibration
    technique 1s used, add  10  uL of  Internal standard to the sample prior to
    injection.

         7.4.2   Follow Section 7.6   1n  Method  8000  for Instructions on the
    analysis sequence,  appropriate  dilutions,  establishing dally retention
    time windows,  and Identification criteria.   Include a mid-level standard
    after  each  group of 10  samples 1n the analysis sequence.

         7.4.3   Examples of GC/ECD  chromatograms  for  phthalate esters are
    shown  in Figures 1 and  2.

         7.4.4   Record the  sample   volume  Injected  and  the  resulting peak
    sizes  (in area units or peak heights).

         7.4.5   Using either the  internal  or external calibration procedure
     (Method 8000),  determine the Identity and  quantity of each analyte peak
    1n  the  sample chromatogram.    See Section  7.8  of  Method  8000 for
    calculation equations.

         7.4.6   If peak detection  and   Identification  are  prevented due to
     interferences,  the hexane  extract may undergo cleanup using either Method
     3610 or 3620.

     7.5  Cleanup;

         7.5.1   Proceed with   either Method   3610  or  3620,   using the 2-mL
     hexane extracts obtained  from  Paragraph  7.1.2.5.

          7.5.2    Following cleanup,  the extracts  should be  analyzed by GC, as
     described  in the previous  paragraphs and  in Method  8000.


8.0  QUALITY CONTROL

     8.1  Refer  to  Chapter  One  for  specific   quality   control  procedures.
Quality control  to validate sample  extraction 1s covered 1n  Method 3500  and  in
the extraction  method utilized.  If  extract cleanup was performed,  follow  the
QC in Method 3600 and 1n the specific cleanup method.

     8.2  Procedures to check  the   GC  system  operation   are found in  Method
8000,  Section 8.6.

          8.2.1   The quality control   check  sample  concentrate (Method  8000,
     Section 8.6) should contain  each  analyte of  interest at the following
     concentrations in acetone:    butyl  benzyl   phthalate,   10 ug/mL;  bis(2-
     ethylhexyl) phthalate, 50 ug/mL; di-n-octyl   phthalate,  50 ug/mL;  and  any
     other phthalate, 25 ug/mL.
                                  8060 - 7
                                                         Revision
                                                         Date  September 1986

-------
           Column:  1.5%SP-2250+
                   1.95% SP-2401 on Supticoport
           Ttmperaturt: 180°C
           Dttictor: Electron Capture
           j    t
                         J	t
      0    2    4    6    8    10   12

         RETENTION TIME (MINUTES)


Figure 1. Gas chromatogram of phthalates (example 1).
         8060 - 8
                                  Revision       0
                                  Date  September 1986

-------
     Column:  1.5% SP-2250+
             1.95% SP-2401 on Supelcoport
     Temperature: 18(X>C
     Detector: Electron Capture
           4       8       12       16

           RETENTION TIME (MINUTES)
18
Figure 2. Gas chromatogram of phthalates (example 2).
                    8060  - 9
                                             Revision      0
                                             Date   September  1986

-------
          8.2.2  Table 3 indicates the  calibration  and  QC  acceptance  criteria
     for this  method.    Table  4  gives  method accuracy  and   precision  as
     functions of concentration for the analytes of  interest.   The contents  of
     both Tables should be used to  evaluate a laboratory's ability to perform
     and generate acceptable data by this method.

     8.3  Calculate surrogate standard  recovery  on  all  samples, blanks,  and
spikes.  Determine if  the  recovery  is  within limits  (limits established  by
performing QC procedures outlined in Method 8000, Section  8.10).

          8.3.1  If recovery is not within limits, the following is required.

                •  Check to  be  sure  there  are  no  errors  in calculations,
                  surrogate solutions  and  internal  standards.   Also, check
                  instrument performance.

                •  Recalculate the data and/or reanalyze  the extract if any of
                  the above checks reveal a problem.

                •  Reextract and reanalyze the sample  if none of the above are
                  a problem or flag the data as  "estimated concentration."


9.0  METHOD  PERFORMANCE

     9.1  The method  was  tested  by  16   laboratories  using  reagent water,
drinking water,  surface water, and  three  industrial wastewaters  spiked at six
concentrations  over the range  0.7  to   106  ug/L.  Single operator precision,
overall precision, and method accuracy  were  found  to be directly related to
the concentration of  the  analyte  and   essentially  independent  of the sample
matrix.    Linear  equations  to  describe  these  relationships   for  a flame
ionization detector are presented in  Table  4.

     9.2  The accuracy  and precision  obtained will be determined  by the sample
matrix, sample-preparation technique,  and  calibration procedures  used.


10.0   REFERENCES

1.  Development and Application of Test   Procedures for Specific  Organic Toxic
Substances in Wastewaters.   Category  1   -  Phthalates.  Report  for EPA  Contract
68-03-2606  (in  preparation).

2.   "Determination of  Phthalates  in  Industrial   and Municipal  Wastewaters,"
EPA-600/4-81-063,    U.S.    Environmental    Protection   Agency,   Environmental
Monitoring and  Support  Laboratory,  Cincinnati,  Ohio 45268,  October 1981.

3.   Burke, J.A.  "Gas  Chromatography  for  Pesticide   Residue Analysis; Some
Practical  Aspects,"   Journal   of   the  Association  of   Official Analytical
Chemists,  48,  1037,  1965.
                                   8060 - 10
                                                          Revision      0
                                                          Date  September 1986

-------
4.  "EPA Method Validation Study  16,  Method  606 (Phthalate Esters)," Report
for EPA Contract 68-03-2606 (in preparation).

5.  U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
Analysis of Pollutants Under the Clean Water Act; Final Rule and Interim Final
Rule and Proposed Rule," October 26, 1984.

6.  Provost, L.P. and R.S.  Elder,   "Interpretation of Percent Recovery Data,"
American Laboratory, 15, pp. 58-63,  1983.
                                   8060  -  11
                                                          Revision
                                                          Date   September 1986

-------
TABLE 3.  QC ACCEPTANCE CRITERIA3


Parameter
Bi s (2-ethy 1 hexy 1 ) phthal ate
Butyl benzyl phthal ate
Di-n-butyl phthal ate
Di ethyl phthal ate
Dimethyl phthal ate
Di-n-octyl phthal ate
Test
cone.
(ug/L)
50
10
25
25
25
50
Limit
for s
(ug/L)
38.4
4.2
8.9
9.0
9.5
13.4
Range
for 7
(ug/L)
1.2-55.9
5.7-11.0
10.3-29.6
1.9-33.4
1.3-35.5
D-50.0
Range
P, PS
(%)
D-158
30-136
23-136
D-149
D-156
D-114
     s = Standard deviation of four recovery measurements, in ug/L.

     7 = Average recovery for four recovery measurements, in ug/L.

     P, Ps = Percent recovery measured.

     D = Detected; result must be greater than zero.

     Criteria from 40 CFR  Part 136 for  Method 606.  These criteria are based
directly upon the method performance  data  in  Table 4.  Where necessary, the
limits for recovery have been broadened  to assure applicability of the limits
to  concentrations below those used to develop Table 4.
                                   8060 - 12
                                                          Revision
                                                          Date   September 1986

-------
TABLE 4.  METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION3
Parameter
Bis(2-ethylhexyl) phthalate
Butyl benzyl phthalate
Di-n-butyl phthalate
Di ethyl phthalate
Dimethyl phthalate
Di-n-octyl phthalate
Accuracy, as
recovery, x1
(ug/L)
0.53C+2.02
0.82C+0.13
0.79C+0.17
0.70C+0.13
0.73C+0.17
0.35C-0.71
Single analyst
precision, sr'
(ug/L)
0.807-2.56
0.267+0.04
0.237+0.20
0.277+0.05
0.267+0.14
0.387+0.71
Overall
precision,
S1 (ug/L)
0.737-0.17
0.257+0.07
0.297+0.06
0.457+0.11
0.447+0.31
0.627+0.34
     x1  = Expected  recovery  for  one  or  more  measurements  of  a  sample
           containing a concentration of C, in ug/L.
     sr' = Expected single analyst  standard  deviation  of measurements at an
           average concentration of 7, in ug/L.
     S1  = Expected interlaboratory standard  deviation  of measurements at an
           average concentration found of 7, in ug/L.
     C   = True value for the concentration, in ug/L.
     7   = Average recovery found for measurements of samples containing a
           concentration of C, in ug/L.
     Criteria from 40 CFR Part 136 for Method 606.
                                  8060 - 13
                                                         Revision      0
                                                         Date  September 1986

-------
                                          METHOD 8O60

                                       PHTHALATE ESTERS
(      Start      J

7.1.




(see

ll
—> Choose
appropriate
extraction
procedure
Chapter 2)
    o
                                                     7.4
                                                       Perform GC
                                                     analysis (see
                                                      Method 8000)
 7.1.21
        Exchange
        extract-
  ion solvent to
        hexane
    during micro
  K-O procedures
  7.2
     Set gas
  chromatography
   conditions
 Is IdentlfIce-^S.
tlon C detection
                             Cleanup
                           using Method
                          361O or 362O)
7
	 	 1 Refer to
Method BOOO
for proper
calibration
techniques
                           7.3.2
                              of
                                 Process
                                 a aeries
                                 standards
                           through cleanup
                              procedure:
                             analyze by GC
                                     8060 -  14
                                                                Revision       0
                                                                Date   September 1986

-------
                                  METHOD 8061

               PHTHALATE ESTERS BY CAPILLARY GAS CHROMATOGRAPHY
                   WITH ELECTRON CAPTURE DETECTION (GC/ECD)
1.0   SCOPE AND APPLICATION

      1.1   Method 8061 is used to determine the identities and concentrations
of various phthalate esters  in liquid, solid and sludge matrices.  The following
compounds can be determined by this method:
      Compound Name                                           CAS No."


      Benzyl benzoate  (I.S.)                                 120-51-4
      Bis(2-ethylhexyl) phthalate                             117-81-7
      Butyl benzyl phthalate                                   85-68-7
      Di-n-butyl phthalate                                     84-74-2
      Diethyl phthalate                                        84-66-2
      Dimethyl phthalate                                      131-11-3
      Di-n-octyl phthalate                                    117-84-0


      a  Chemical  Abstract Services Registry Number.

      1.2   Table  1  lists the  method  detection  limits  (MDL)  for  the target
analytes in a water matrix.  The MDLs for the components of a specific sample may
differ  from those listed in Table  1  because  MDLs  depend  on  the  nature  of
interferences in the  sample  matrix.  Table  2 lists the estimated quantisation
limits (EQL) for other matrices.

      1.3   When this method is used to  analyze for any  or  all  of the target
analytes, compound identification should be supported by at least one additional
qualitative technique.  This method  describes  conditions for parallel column,
dual electron capture detector  analysis which  fulfills the above requirement.
Retention time  information obtained  on two  megabore  fused-silica open tubular
columns is given in Table  1.  Alternatively, gas chromatography/mass spectrometry
could be used for compound confirmation.

      1.4   The  following compounds,  bis(2-n-butoxyethyl)  phthalate,  bis(2-
ethoxyethyl) phthalate, bis(2-methoxyethyl)  phthalate, bis(4-methyl-2-pentyl)
phthalate,   diamyl  phthalate,   dicyclohexyl   phthalate,   dihexyl   phthalate,
diisobutyl  phthalate, dinonyl phthalate,  and hexyl  2-ethylhexyl  phthalate can
also be analyzed by this method and may be used as surrogates.

      1.5   This method is  restricted  to use  by  or under the  supervision  of
analysts experienced  in  the use  of gas chromatographs  and  skilled in  the
interpretation of gas  chromatograms.  Each analyst must demonstrate the ability
to generate acceptable results with this  method.
                                   8061 - 1                         Revision 0
                                                                September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   A measured volume  or  weight of sample  (approximately  1  liter for
liquids,  10  to 30  grams  for  solids  and sludges)  is  extracted by  using the
appropriate sample extraction technique specified in Methods 3510,  3540, 3541,
and 3550.  Method 3520 is  not recommended for the extraction of aqueous samples
because the longer chain esters (dihexyl  phthalate, bis(2-ethylhexyl) phthalate,
di-n-octyl phthalate,  and  dinonyl phthalate) tend to  adsorb to the glassware and
consequently, their extraction  recoveries are <40 percent.  Aqueous samples are
extracted at a  pH of  5 to 7,  with methylene chloride,  in a separatory funnel
(Method 3510).  Alternatively,  particulate-free aqueous samples could be filtered
through membrane disks that  contain C18-bonded  silica.  The phthalate esters are
retained by the silica and,  later eluted with  acetonitrile.  Solid samples are
extracted with  hexane/acetone  (1:1)  or methylene chloride/acetone  (1:1)  in a
Soxhlet extractor  (Methods  3540/3541) or with  an  ultrasonic extractor (Method
3550).   After  cleanup,  the extract  is analyzed by  gas  chromatography  with
electron capture detection  (GC/ECD).

      2.2   The  sensitivity of Method  8061  usually depends  on the  level  of
interferences rather than  on instrumental limitations.   If  interferences prevent
detection of the analytes, cleanup of the sample extracts  is necessary.  Either
Method 3610 or  3620 alone or  followed  by  Method 3660,  Sulfur Cleanup, may be used
to eliminate  interferences in the analysis.  Method 3640,  Gel  Permeation Cleanup,
is applicable for samples that contain high amounts  of lipids and waxes.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000.

      3.2   Interferences coextracted from the  samples  will vary considerably
from waste to waste.  While general cleanup techniques are referenced or provided
as part of this method, unique samples may require additional cleanup approaches
to achieve desired sensitivities for the target analytes.

      3.3   Glassware  must  be scrupulously  clean.    All  glassware  require
treatment in a muffle  furnace at 400  °C for 2 to  4  hrs, or  thorough rinsing with
pesticide-grade  solvent,  prior to use.  Refer  to Chapter  4,  Sec.  4.1.4, for
further details regarding  the cleaning of glassware.  Volumetric glassware should
not be heated  in  a muffle furnace.

      If Soxhlet extractors  are baked in the muffle  furnace, care must be taken
to ensure  that they  are  dry (breakage may result if any  water  is  left in the
side-arm).  Thorough rinsing with hot tap water,  followed by deionized water and
acetone  is  not an adequate decontamination procedure.    Even  after  a Soxhlet
extractor was refluxed with  acetone for three days,  with daily solvent changes,
the concentrations  of bis(2-ethylhexyl) phthalate were as  high  as 500 ng per
washing.  Storage of glassware in  the laboratory introduces  contamination, even
if the glassware  is wrapped  in  aluminum foil.  Therefore,  any glassware used in
Method 8061 should be cleaned  immediately prior to use.

      3.4   Florisil and alumina may  be contaminated with  phthalate esters and,
therefore,  use  of these  materials  in  sample  cleanup   should   be  employed


                                   8061  - 2                         Revision 0
                                                                September 1994

-------
cautiously.  If these materials are used,  they must be obtained packaged  in glass
(plastic  packaging  will contribute  to contamination with  phthalate esters).
Washing of  these  materials prior to use with  the  solvent(s)  used for elution
during extract cleanup was  found helpful,  however, heating at 320 °C for Florisil
and  210  CC  for  alumina is  recommended.    Phthalate  esters were  detected in
Florisil  cartridge method  blanks  at  concentrations  ranging  from 10 to 460 ng,
with 5 phthalate  esters  in the 105 to 460 ng  range.  Complete removal  of the
phthalate esters  from  Florisil cartridges  does  not seem possible,  and  it is
therefore  desirable  to  keep the  steps  involved in  sample preparation  to  a
minimum.

      3.5    Paper thimbles and filter paper must be  exhaustively washed with the
solvent that will  be used in  the sample extraction.   Soxhlet  extraction of paper
thimbles and filter paper  for  12 hrs with fresh solvent should be repeated for
a minimum of three times.  Method  blanks  should  be  obtained before any of the
precleaned thimbles or filter papers are used.   Storage of precleaned thimbles
and  filter  paper  in  precleaned  glass  jars  covered  with  aluminum  foil  is
recommended.

      3.6    Glass wool  used  in  any step  of  sample  preparation should  be  a
specially treated pyrex  wool,  pesticide  grade, and  must  be  baked at 400°C for
4 hrs. immediately prior to  use.

      3.7    Sodium sulfate must be obtained packaged in glass  (plastic packaging
will contribute to contamination with phthalate esters),  and  must be purified by
heating at 400 °C  for 4  hrs.  in a shallow tray,  or by precleaning with methylene
chloride (Sec. 5.3).  To avoid recontamination, the  precleaned material must be
stored in  glass-stoppered glass bottles, or glass bottles  covered with precleaned
aluminum foil.  The  storage  period should not exceed two weeks.   To minimize
contamination, extracts should be  dried directly  in  the glassware  in which they
are  collected  by  adding small amounts  of  precleaned sodium  sulfate  until  an
excess of free flowing material is noted.

      3.8    The presence of  elemental  sulfur will  result  in large peaks which
often mask  the  region  of the  compounds eluting  before  dicyclohexyl  phthalate
(Compound No. 14) in the gas chromatograms  shown in Figure  1.   Method 3660 is
suggested for removal  of sulfur.

      3.9   Waxes and lipids can  be removed by  Gel  Permeation Chromatography
(Method 3640).  Extracts containing high concentrations  of lipids are viscous,
and may even solidify at room temperature.


4.0   APPARATUS AND MATERIALS

      4.1    Gas chromatography

            4.1.1 Gas  chromatograph,  analytical  system  complete  with  gas
      chromatograph suitable  for on-column  and split/splitless injections and
      all   required  accessories,   including   detector,   analytical   columns,
      recorder, gases,  and syringes.  A data system for  measuring peak heights
      and/or peak areas is recommended.
                                   8061 - 3                         Revision 0
                                                                September 1994

-------
                  4.1.1.1     Eight inch injection tee (Supelco, Inc., Catalog
            No. 2-3665, or equivalent)  or glass  Y  splitter for megabore columns
            (J&W Scientific, -"press-fit",  Catalog  No. 705-0733, or equivalent).

            4.1.2 Columns

                  4.1.2.1     Column 1, 30 m x 0.53 mm  ID, 5% phenyl/95% methyl
            silicone fused-silica open  tubular column (DB-5, J&W Scientific, or
            equivalent), 1.5 /xm film thickness.

                  4.1.2.2     Column 2,  30  m x  0.53  mm ID,  14%  cyanopropyl
            phenyl  silicone  fused-silica  open  tubular  column  (DB-1701,  J&W
            Scientific, or equivalent), 1,0 ptm  film thickness.

            4.1.3 Detector - Dual electron capture detector (ECD)

      4.2   Glassware, see Methods 3510, 3540, 3541,  3550, 3610, 3620, 3640, and
3660 for specifications.

      4.3   Kuderna-Danish (K-D) apparatus.

            4.3.1 Concentrator tube - 10 ml graduated  (Kontes K-570050-1025 or
      equivalent).  A  ground  glass  stopper  is  used to prevent evaporation of
      extracts.

            4.3.2 Evaporation flask  -  500 ml (Kontes  K-570001-500  or equiva-
      lent).  Attach to concentrator tube with springs, clamps, or equivalent.

            4.3.3 Snyder  column  -  Three  ball macro (Kontes  K-503000-0121 or
      equivalent).

            4.3.4 Snyder  column  -  Two  ball  micro  (Kontes K-569001-0219  or
      equivalent).

            4.3.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).

      4.4   Boiling chips, approximately 10/40 mesh.  Heat to 400  °C for 30 min,
or Soxhlet-extract with methylene chloride prior to use.

      4.5   Water  bath,  heated,  with  concentric  ring  cover,  capable  of
temperature control (+ 2°C).


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests.  Unless otherwise
indicated, it is intended that all reagents shall  conform to the specifications
of the Committee on Analytical  Reagents of the American Chemical Society, where
such specifications are  available.   Other grades may  be  used,  provided it is
first ascertained that the reagent  is of sufficiently high purity to permit its
use without lessening the accuracy of the determination.
                                   8061 - 4                         Revision 0
                                                                September 1994

-------
      5.2   Organic-free reagent water.  All  references  to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Sodium sulfate (granular, anhydrous),  Na2S04.  Purify by heating at
400 "C for 4 hours in a shallow tray, or  by precleaning  the sodium sulfate with
methylene chloride.   If the sodium sulfate is precleaned with methylene chloride,
a method blank must be analyzed, demonstrating that there is no interference from
the sodium sulfate.

      5.4   Solvents:

            5.4.1 Hexane, C6H14  -  Pesticide quality,  or  equivalent.

            5.4.2 Methylene chloride, CH2C12 - Pesticide  quality, or equivalent.

            5.4.3 Acetone, CH3COCH3 - Pesticide  quality, or equivalent.

            5.4.4 Acetonitrile, CH3CN - HPLC  grade.

            5.4.5 Methanol, CH3OH - HPLC  grade.

            5.4.6 Diethyl Ether,  C2H5OC2H5  -  Pesticide quality, or equivalent.
      Must be free  of  peroxides,  as indicated by  test strips  (EM  Quant,  or
      equivalent).   Procedures  for removal of  peroxides are  provided with the
      test strips.  After cleanup, 20 ml of ethyl  alcohol preservative must be
      added to each liter of ether.

      5.5   Stock standard solutions:

            5.5.1 Prepare  stock  standard  solutions at  a  concentration  of
      1000 mg/L by dissolving 0.0100 g  of assayed reference material in hexane,
      and diluting to volume in a 10 ml volumetric flask.  When compound purity
      is assayed  to  be  96 percent or greater,  the weight can be used without
      correction  to  calculate  the  concentration  of  the  stock  standard.
      Commercially  prepared  stock  standard  solutions  can   be  used  at  any
      concentration  if  they  are certified  by  the manufacturer  or  by  an
      independent source.

            5.5.2 Transfer the  stock standard  solutions into glass  vials with
      Teflon  lined screw-caps  or crimp tops.   Store  at 4 °C and protect from
      light.   Stock  standard solutions should  be checked periodically  by gas
      chromatography for signs  of degradation  or  evaporation, especially just
      prior to preparation of calibration standards.

            5.5.3 Stock standard solutions must be replaced  after 6 months,  or
      sooner  if comparison with check standards indicates a  problem.

      5.6   Calibration  standards:  Calibration standards  are  prepared  at  a
minimum of five concentrations for each parameter  of interest through dilution
of the stock  standard solutions with hexane.   One  of the concentrations should
be at  a concentration near, but above, the method detection limit.  The remaining
concentrations should correspond to the expected range of concentrations  found
in real  samples,  or  should define  the working  range of the  GC.   Calibration


                                   8061 - 5                         Revision 0
                                                                September 1994

-------
solutions must be  replaced  after 1 to 2 months,  or  sooner if comparison with
calibration verification standards indicates a problem.

      5.7   Internal standards (if internal standard calibration is used):  To
use this approach, the analyst must select one or more internal standards that
are similar in analytical behavior to  the  compounds  of interest.   The analyst
must further demonstrate that the  measurement  of the internal  standard is not
affected by method or matrix interferences.  Benzyl benzoate has been tested and
found appropriate for Method 8061.

            5.7.1 Prepare a spiking  solution  of benzyl benzoate  in  hexane at
      5000 mg/L.   Addition of 10 juL of this solution to 1 ml of sample extract
      is recommended.  The spiking concentration of the  internal  standard should
      be kept constant  for  all  samples and calibration standards.   Store the
      internal  standard  spiking  solution at 4  °C in glass vials  with Teflon
      lined screw-caps  or crimp  tops.   Standard  solutions  should  be replaced
      when ongoing QC (Sec.  8) indicates a problem.

      5.8   Surrogate standards:  The analyst should monitor the performance of
the extraction, cleanup  (when used), analytical  system,  and  the  effectiveness of
the method in dealing  with each sample  matrix by  spiking each sample, standard,
and blank with surrogate compounds. Three surrogates may be used for Method 8061
in  addition to   those   listed   in  Sec.   1.4:   diphenyl  phthalate,  diphenyl
isophthalate, and dibenzyl phthalate.  However,  the compounds listed in Sec. 1.4
are recommended.

            5.8.1 Prepare a surrogate  standard  spiking solution,  in acetone,
      which contains  50  ng//xL of each compound.   Addition of  500 juL  of this
      solution  to 1 L  of water or 30 g  solid sample is equivalent to 25 fj.g/1 of
      water or 830  jug/kg of  solid sample.   The  spiking concentration  of the
      surrogate standards may be  adjusted  accordingly,  if  the final  volume of
      extract   is  reduced below  2 ml  for  water  samples   or  10 ml  for  solid
      samples.  Store  the surrogate spiking solution at 4 °C in  glass vials with
      Teflon lined  screw-caps  or crimp tops.   The solution must  be replaced
      after 6  months,  or sooner if ongoing QC (Sec. 8) indicates problems.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material  to this chapter, Organic  Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Extraction:

            7.1.1 Refer to Chapter Two for  guidance on choosing the appropriate
      extraction procedure.  In general,  water  samples are extracted at a pH of
      5 to  7  with methylene  chloride in  a  separatory  funnel  (Method  3510).
      Method  3520  is  not  recommended for the  extraction of  aqueous  samples
      because  the  longer  chain  esters  (dihexyl  phthalate  bis(2-ethylhexyl)
      phthalate, di-n-octyl phthalate, and  dinonyl phthalate) tend to adsorb to


                                   8061 - 6                         Revision 0
                                                                September 1994

-------
the  glassware   and  consequently,  their   extraction   recoveries  are
<40 percent.  Solid samples are extracted with  hexane/acetone (1:1) or
methylene  chloride/acetone   (1:1)   in   a   Soxhlet  extractor  (Methods
3540/3541) or with  an ultrasonic extractor  (Method  3550).   Immediately
prior  to  extraction,  spike 500 /uL  of  the  surrogate  standard  spiking
solution (concentration = 50 ng//iL) into 1  L  aqueous  sample or 30 g solid
sample.

      7.1.2 Extraction   of  particulate-free   aqueous    samples   using
C18-extraction disks (optional):

            7.1.2.1     Disk preconditioning: Place the C18-extraction disk
      into the filtration apparatus and  prewash the disk with 10 to 20 ml
      of acetonitrile.  Apply vacuum to pull the solvent  through the disk.
      Maintain vacuum to pull air through for 5 min.  Follow with 10 ml of
      methanol.   Apply  vacuum and  pull  most  of the methanol  through the
      disk.  Release vacuum before the disk gets  dry.  Follow with 10 ml
      organic-free reagent water.  Apply vacuum and pull  most of the water
      through the disk.  Release the vacuum before the disk gets dry.

            7.1.2.2     Sample preconcentration:  Add 2.5 ml of methanol to
      the  500 ml  aqueous sample in  order  to get  reproducible results.
      Pour the  sample into the filtration apparatus.   Adjust  vacuum so
      that it takes approximately  20 min to  process  the  entire sample.
      After all  of the sample has passed through  the membrane disk, pull
      air through the disk for  5  to  10 min. to remove any residual  water.

            7.1.2.3     Sample elution:  Break the vacuum and place the tip
      of the filter base into the test tube that  is contained inside the
      suction flask.  Add 10 ml of acetonitrile to the graduated funnel,
      making sure  to  rinse the walls of the graduated funnel  with  the
      solvent.  Apply  vacuum to pass the  acetonitrile  through the membrane
      disk.

            7.1.2.4     Extract concentration (if necessary): Concentrate
      the extract to  2 mL  or less,  using either  the  micro Snyder  column
      technique  (Sec.  7.1.2.4.1)  or  nitrogen blowdown technique  (Sec.
      7.1.2.4.2).

                  7.1.2.4.1   Micro Snyder Column Technique

                        7.1.2.4.1.1 Add one or two  clean boiling chips to
                  the concentrator  tube and attach a two  ball micro Snyder
                  column.   Prewet  the  column by adding about  0.5  mL of
                  acetonitrile to the top  of  the column.   Place the  K-D
                  apparatus in  a hot water bath (15-20°C  above the boiling
                  point of the solvent)  so that the concentrator tube is
                  partially immersed in the hot water  and the entire lower
                  rounded surface of the flask is bathed with hot  vapor.
                  Adjust the vertical position of  the  apparatus  and  the
                  water  temperature,  as  required,  to   complete  the
                  concentration in 5-10 minutes.    At  the  proper rate of
                  distillation the  balls  of .the  column  will  actively


                             8061 - 7                         Revision 0
                                                          September 1994

-------
                        chatter, but  the  chambers will  not  flood.   When  the
                        apparent volume of liquid reaches 0.5  ml, remove the K-D
                        apparatus from the water  bath and allow it to drain and
                        cool for at least 10 minutes.  Remove the Snyder column
                        and rinse  the flask and  its  lower joints  with  about
                        0.2 ml  of  solvent  and add  to  the concentrator  tube.
                        Adjust the final volume to 1.0-2.0 ml with solvent.

                        7.1.2.4.2   Nitrogen Slowdown  Technique

                              7.1.2.4.2.1  Place the  concentrator tube  in a warm
                        water  bath  (approximately  35 °C)  and  evaporate  the
                        solvent volume  to  the required  level  using  a  gentle
                        stream of clean, dry nitrogen (filtered through a column
                        of activated carbon).

                              CAUTION:    Do not  use plasticized tubing between
                                          the carbon trap and the sample.

                              7.1.2.4.2.2  The internal  wall of the tube must be
                        rinsed down several  times with  acetonitrile during the
                        operation.   During evaporation,  the solvent level in the
                        tube must be positioned to prevent water from condensing
                        into the sample  (i.e., the solvent level should be below
                        the level of  the water bath).   Under normal operating
                        conditions, the extract  should  not be allowed to become
                        dry.

      7.2   Solvent Exchange: Prior to Florisil  cleanup or gas chromatographic
analysis, the methylene chloride and methylene chloride/acetone extracts obtained
in Sec. 7.1.1  must be exchanged to hexane,  as described in Sees.  7.2.1 through
7.2.3.   Exchange is  not  required   for  the  acetonitrile extracts  obtained in
Sec. 7.1.2.4.

            7.2.1 Add one or two clean boiling chips to the flask and attach a
      three ball Snyder column.  Concentrate the  extract  as  described in  Sec.
      7.1.2.4.1, using  1  ml of methylene chloride  to  prewet the  column,  and
      completing the concentration in 10-20  minutes.   When the apparent volume
      of liquid reaches  1-2  ml,  remove the  K-D apparatus from  the water bath and
      allow it to drain and cool for at least 10 minutes.

            7.2.2 Momentarily remove the Snyder column, add 50 ml of hexane, a
      new boiling chip,  and attach the macro Snyder column.   Concentrate the
      extract as described in  Sec. 7.1.2.4.1, using 1 ml of hexane to prewet the
      Snyder column,  raising the temperature of the water bath, if necessary, to
      maintain proper distillation, and completing the  concentration  in  10-20
      minutes.  When  the  apparent  volume of liquid  reaches  1-2  ml,  remove the
      K-D apparatus and allow it to drain  and cool for  at least 10 min.

            7.2.3 Remove  the  Snyder column  and  rinse  the flask  and its  lower
      joint into the  concentrator tube with 1 to 2 ml hexane.  A 5 ml syringe is
      recommended for this  operation.   Adjust  the extract volume  to  2 ml for
      water samples,  using  either the micro  Snyder  column technique  (Sec.


                                   8061 -  8                          Revision 0
                                                                September 1994

-------
7.1.2.4.1) or nitrogen blowdown technique (Sec. 7.1.2.4.2), or 10 mL for
solid samples.  Stopper the concentrator tube and store at 4 °C  if further
processing will be performed immediately.  If the extract will be stored
for two days  or longer,  it should  be transferred  to a glass vial with a
Teflon lined screw-cap or crimp top.  Proceed with  the gas chromatographic
analysis.

7.3   Cleanup/Fractionation:

      7.3.1 Cleanup may  not  be  necessary  for extracts from a relatively
clean   sample  matrix.    If   polychlorinated   biphenyls   (PCBs)   and
organochlorine pesticides are known to be present in the sample, use the
procedure outlined in  Methods 3610  or 3620.   When using column cleanup,
collect  Fraction   1  by  eluting with 140 ml  (Method  3610)   or  100 ml
(Method 3620) of  20-percent  diethyl ether in hexane.   Note that,  under
these   conditions,   bis(2-methoxyethyl)   phthalate,  bis(2-ethoxyethyl)
phthalate, and bis(2-n-butoxyethyl) phthalate are not recovered from the
Florisil column.  The elution patterns and compound recoveries are given
in Table 3.

      7.3.2 Methods 3610 and 3620  also describe   procedures  for sample
cleanup  using  Alumina  and  Florisil  Cartridges.    With   this  method,
bis(2-methoxyethyl)   phthalate,   bis(2-ethoxyethyl)   phthalate,    and
bis(2-n-butoxyethyl) phthalate are recovered quantitatively.

7.4   Gas chromatographic conditions  (recommended):

      7.4.1 Column 1 and Column 2 (Sec. 4.1.2):

      Carrier gas (He) =            6 mL/min.
      Injector temperature =        250 "C.
      Detector temperature =         320 °C.
      Column temperature:
            Initial temperature =   150 °C,  hold for 0.5 min.
            Temperature  program =   150  °C   to 220  °C  at  5  °C/min.,
                                    followed by  220 °C  to 275 °C  at  3
                                    "C/min.
            Final  temperature =     275 °C hold for 13 min.

      7.4.2 Table  1  gives the  retention  times  and  MDLs  that can  be
achieved by this  method  for  the 16  phthalate esters.   An example of the
separations achieved with the DB-5 and DB-1701 fused-silica open tubular
columns is shown  in Figure 1.

7.5   Calibration:

      7.5.1 Refer to Method 8000 for proper  calibration techniques.   Use
Tables  1  and  2  for guidance  on   selecting the  lowest  point  on  the
calibration curve.

      7.5.2 The procedure  for  internal or  external calibration may  be
used.    Refer  to   Method 8000  for  the  description  of  each  of  these
procedures.
                             8061 - 9                         Revision 0
                                                          September 1994

-------
      7.6   Gas chromatographic analysis:

            7.6.1 Refer to Method 8000.   If  the internal  standard calibration
      technique is used, add  10 juL  of internal  standard solution at 5000 mg/L
      to the sample prior to injection.

            7.6.2 Follow Method 8000 for instructions on the analysis sequence,
      appropriate dilutions,  establishing daily  retention  time  windows,  and
      identification criteria.

            7.6.3 Record the  sample  volume  injected  and  the  resulting  peak
      areas.

            7.6.4 Using  either  the  internal  or  the  external  calibration
      procedure (Method 8000), determine the identity and the quantity of each
      component  peak  in  the  sample chromatogram  which  corresponds  to  the
      compounds used for calibration purposes.

            7.6.5 If the response of a peak exceeds the working  range of the
      system, dilute the extract and reanalyze.

            7.6.6 Identify compounds  in  the  sample  by  comparing the retention
      times of the peaks in the sample chromatogram with those of the peaks in
      standard  chromatograms.    The  retention  time  window  used  to  make
      identifications  is  based  upon measurements  of  actual   retention  time
      variations over the course of  10 consecutive injections.  Three times the
      standard  deviation  of  the  retention  time  can  be  used  to  calculate  a
      suggested window size.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter  One  for   specific  quality control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was  performed,  follow the QC
specified in Method 3600 and in the specific cleanup method.

      8.2   Quality control  required to evaluate  the GC system operation  is found
in Method 8000.

            8.2.1 The  quality  control  check sample concentrate  (Method  8000)
      should contain the test compounds  at 5 to 10 ng/juL.

      8.3   Calculate the recoveries  of the surrogate compounds  for all samples,
method blanks, and method spikes.  Determine if the recoveries are within limits
established by performing QC procedures  outlined in Method 8000.

            8.3.1 If the recoveries  are not within limits,  the following are
      required:

                  8.3.1.1     Make sure  there are no  errors  in calculations,
            surrogate  solutions and internal standards.  Also check instrument
            performance.


                                  8061  - 10                         Revision 0
                                                                September 1994

-------
                  8.3.1.2     Recalculate the data and/or reanalyze the extract
            if any of the above checks reveal a problem.

                  8.3.1.3     Reextract and reanalyze the sample  if none of the
            above are a problem,  or flag  the  data  as  "estimated concentration."

      8.4   An  internal  standard  peak area  check must  be performed  on all
samples.  The internal  standard must be evaluated  for acceptance  by determining
whether the measured  area  for the  internal  standard  deviates  by more than 30
percent from  the average  area for  the  internal  standard  in  the calibration
standards.  When the internal standard  peak area is outside  that  limit, all
samples that fall outside the QC criteria must be reanalyzed.

      8.5   GC/MS confirmation: Any compounds confirmed  by two columns may also
be confirmed by GC/MS if the concentration  is sufficient  for detection by GC/MS
as determined by the laboratory-generated detection limits.

            8.5.1 The GC/MS would normally require a minimum concentration  of 10
      ng//iL in the final extract for each single-component  compound.

            8.5.2 The sample extract and  associated blank should  be analyzed by
      GC/MS as per Sec.  7.0  of Method  8270.   Normally,  analysis  of a blank is
      not required for confirmation  analysis, however, analysis for phthalates
      is  a  special  case because of the  possibility  for sample  contamination
      through septum punctures, etc.

            8.5.3 A reference standard  of the compound must  also  be analyzed by
      GC/MS.   The  concentration  of  the  reference  standard  must   be   at  a
      concentration that would demonstrate  the ability to confirm the phthalate
      esters identified by GC/ECD.

      8.6   Include a mid-concentration calibration standard after each group of
20 samples  in   the   analysis  sequence.     The   response   factors   for   the
mid-concentration calibration must  be within  + 15  percent of the average values
for the multiconcentration calibration.

      8.7   Demonstrate  through  the analyses of  standards that  the  Florisil
fractionation scheme  is reproducible. When using the fractionation schemes given
in Methods 3610  or 3620,  batch-to-batch  variations  in the  composition  of the
alumina or  Florisil  material may  cause  variations  in  the recoveries  of the
phthalate esters.


9.0   METHOD PERFORMANCE

      9.1    The MDL is defined in Chapter One. The MDL concentrations listed in
Table 1 were  obtained  using  organic-free  reagent water.   Details  on  how  to
determine MDLs are given in Chapter One.   The MDL actually achieved in a given
analysis will vary,  as  it  is  dependent  on  instrument sensitivity and  matrix
effects.

      9.2    This method has been tested  in a single laboratory by using different
types of aqueous samples and  solid  samples which  were fortified  with  the  test


                                   8061 - 11                         Revision  0
                                                                September  1994

-------
compounds at two concentrations.  Single-operator precision, overall precision,
and method  accuracy  were  found to  be  related  to  the  concentration  of the
compounds and  the type  of  matrix.    Results  of the  single-laboratory method
evaluation are presented in Tables 4 and 5.

      9.3   The  accuracy  and  precision  obtained  is  determined by  the sample
matrix,  sample  preparation  technique,  cleanup techniques,   and  calibration
procedures used.


10.0  REFERENCES

1.    Glazer, J.A.;  Foerst, G.D.;  McKee, G.D.; Quave, S.A.,  and  Budde, W.L.,
      "Trace Analyses  for  Wastewaters,"  Environ.  Sci. and  Technol.  15: 1426,
      1981.

2.    Lopez-Avila, V.,  Baldin,  E.,  Benedicto,  J.,  Milanes,  J.,  and Beckert,
      W.F.,  "Application of Open-Tubular Columns to SW-846 GC Methods", EMSL-Las
      Vegas, 1990.

3.    Beckert, W.F. and Lopez-Avila, V., "Evaluation of SW-846 Method 8060 for
      Phthalate  Esters",  Proceedings  of  Fifth  Annual   Testing   and   Quality
      Assurance Symposium, USEPA, 1989.
                                   8061 - 12
    Revision 0
September 1994

-------
O •— i





<0
cS
LU
t—
OO
LU
LU
1—

Q •«- C
y i ^^

CM

CD C
-(-> r—
^^ O
C C 0
0 •!-
•'- E
i ^ ^ -.
c > — i
cu
4-> «=
a: 3

o
o
r— 4-> >,
fO L> i-
0 03 4->
•r- i_  •
£ 4-> -r- 0
CU (/) C7>Z
-C J2 OJ
o  f^ f""^
^i~ LO CM CO f^^
IO CM r— I CO CO



1 — LO r— 1 IO 1 —
co ^r en vo CM
10 oo CM ^- 10






to ^"^ ^^~ tO r*1**
C5 CO ^^ C\J f**+
h^ O^ ^i- to 00
r-H i—H i— 1


co CM 10 CM en
i i i i i
r"H *O Cn ^^ CD
i— i 10 10 r~- LO
i i i i i
»~H ^" <^* ^f" (O
CO 00 00 00 *1-
i-H i — 1







cu
+->
03
r^
03
t~
4->
_C
a.

„ 	 s
p—
QJ QJ >->
4-> +J -(->
OJ OJ O3 C
4-> QJ i— P— QJ
03 4-> 03 OJ Q.
i — OJ JC *"" I
03 i — 4-> -t-> CM
-C 03 x: -c i
-I-J -C Q- CLr—
r~ [ ^ >^
Q.-C r— r— -C
f*> >^ >^ 4^
,— 4J 4J QJ
>»•— 3 3 E
-C >->JD -Q i
4-> ^= O 1 «*•
CU 4-> C^ C ' 	
E QJ -r- i 01
Q Q Q Q CO


I-H CM CO •* LO










O CD
i — i i — i
LO i-H



—< CO

10 00






CM LO
O CM
r^- CD
I-H CM


000
1 1
CM 00
00 — i
1 1
1**^ i-H
I-H CO
i-H i-H










QJ
.(_)
OJ
f—
OJ
.G
+->

a.

^~^
i —
>>
.C QJ
-t-> 4->
CU O3
>>•—
X OJ
O f~
.C +J
4_> f~
CD a.
E
i i —
CM >>

01 OJ
co a


IJD 1*^










CD O
l4^- CO
CM i-H



— I t-«.
CM CT>
00 00






CO f~-
*3" O
cr> •—(
•— I CM


CT> ^1"
l l
^" ^O
LO ^H
1 1
LO CO
o r-
ifi to
LO
^









QJ QJ
-(-> 4->
OJ OJ
r— p—
03 03
-c ^:

^: J=
CL Q.

'^^^r^
T— >^
>> X
J= QJ
•4-^ ^"
CU i —
>> >1

O 4->
-C QJ
4-> 1
QJ CM
I
CM r~
*"""^ ^^

CO 1C


00 
03
r—
03
QJ *"~
+J 4->
OJ -C
t— f*i
OJ
,C i —
+-> >>
-C M
Q. C
QJ
t— XI
>,
X r—
OJ >,

0 CO


0 ^H











03
i— QJ
OS 4->
-C 03
+J r—

fl f—
4-J
*~*~ c-
r— O.
>>

+J r—
QJ >,
-*^ X
X CU
0 -C
1 ' p..
3 ^>
J3 J=
I 4->
C QJ
I i
CM CM
^^^~*
(A 01
00 00


CM CO










CM CTl CM "
CM ^~ CM




LO CO ^i* 1^^
CO 00 00 <— ^
to Crt CO *— H
CM CM fO i—H






00 CO O i — i
00 CO 00 t"»
00 CO 00 CM
CM CO CO I-H


I-.0**- •*
III 1
I-H ^J- »O 1 — 1

OJ OJ
x: r— QJ
+J 03 +J
JZ ^T 03 QJ
<**i 4_3 i— 4->
.C 03 03
r— CL.C 0
>) 4-> ISI
X r— -C C
QJ >i Q. QJ
x: 4-> jQ
O <_J r—
r— 0 >, r-
(J 1 C >,
>, C O M
O i C C
Q Q Q CO


<*• LO UO 00










u





CM
co
00







e~
Q.

j—
~2>^
C
QJ
-C
CL
Q


i — i
i
OO








o u





I-- LO
CO VO
•-H CM
ro co





CT> O
CT> *<•
CM <3-
CO CO


1O CTl
l l
LO •— 1
«d- co
i i
*J- CO
^>- CM
r~~ LO


















QJ
4^
03
•— QJ
O3 4->
J=. OJ
4-> i 	
-C OJ
Q..C
O +->
01 .C.
•r- Q.

f— i>—
>^ >^
C M
QJ C
-C CU
a. .a
Q a


CM CO
=> ID
OO OO


01
•f—
>
QJ
o:































co
i-H
1
i — 1

4->
O.
^
VI








































































-------







































*— X
~o
CO
3
C
•r—
+J
C
o
o
•*~f

•
r— 1

CU
r—
-Q
fO
H-





































CO
(/)
• +-> 3 +J
• •— •» m as
*-** CO CO
CO CO ,—«..,- E CO
£ COT3 3 S_
CO C ,_ s_ .,_ 3
«= -* 0 CO i— +J
-^ O jc +J CO rt>
"'" JC g~ «r— cu
JC +-> -r- i — CO O.
•I-* E Q. as S
_ E co Ol CO
E r— co +->
r™" "r— r— I 4^ <—
•r- <*- ^— Q) Q) i-
**- _ i— T- 0
_ E O C i- -(->
E 3.0 -r- S_ U
3. in nJ co
O 1^ co O +•>
in • CM w cu
• i— i fO "O
i—l •— ' Or— •
• — •*-> O1JC •-
C CLO
C E C -t-> 03 o
E 3 CO -r- S- O
3 r~~ ^ c^. 01 in
r— O +-> O CM
O U CO +J
O •» CO OS CO
S- C CO E S-
S- nj -r- S- o 3
<0 r— E Q. S- +•>
r— 3 \ JC (O
3 J3 O 0 O S-
JO 3 o -r- CU
3 +J in <4- CO Q.
-t-> -i- as E
C +J -l-> Ol CU
C CU OS C -M
CO Q. CU fO
O. O <-J ••- S-
O o o <4- O
«j o co o -i->
OJ O CM U
U •!- CM 3 +J CU
•r- i— e>8 S- -r-j
r— -i- O r3 O C
• r- CO 4J O--i-
co > as
i ^n ^"^ ^— * *^
•o eu -o s- o c
CO co i — O M- •!-
CO 3 O +-> E
3 14- JC CU O \
q_ CU CU — 1
t-i c -t-> -i-j E
in o •"- c
t i-^ E c -r- o
DQ t— i O CM
Q i in -i— cu
CO • +J JC 4->
o a o o +-> ro
i— i *-" CO
Q -r-j O C
E >— ' o c: -M cu
E » ••— Ol
E O co o
co E m o c s-
m <-< o E -(->
• CO 1— 3 -r-
o m co co i — c
• -i- a. o
X O 3 (J CO
E OO as
E E « O Ol
S- C 2
o o avr- -i-> a.
CO CO O I 3
S- CO CO CU
as as *** f~ **s
C +-> as
co co cu «t E
.,- -r- S- +J
3 U •-
,— I CM -M • CU C
OS C C •!-
C C S- "- C E •
E E CU E O^v.0
3 3 Q.^^ (J — 1 o
i— •— E 0 E 0
O O CO o o CM
o o I— ro +j UD co
C .. Q CO
o 5 oo +J
CO -i— •*•* O
| > | ' | Of ^J QJ
(C O T3 T C $-
U as CO *- n> i-
-r- S- CO g- O
r— •«-> 3 ? - O
a. x <4- h E
cu cu g- o ••->
i- --'in S--D o
, « cu c
£= t3 CD -, CU
CU 0 Q 2 S- CO
> JC 3 4- S-
CO +•> r~ CO
co cu •• ™ tt- S
E JC > 0
M- 0 . T3
O r— rO •'-' CO CU
fO O . CO S-
tO 0 S- "^ CU 3
•r- T- a.- j_ co
co -i-> ca.^ 01 i >> as % CU O)
i^— ^— co ^^ E
as as c "2
c c E ^ — • co

r— <" C: 3
CU CO O „. r—
JC S- U & -C as
4-> •!- J= 4-> >
+J CU *•* -r-
E C t— S
O CU Ol ^
S- C -1~ C W
t+» CU *r~ O 4-^
JC CO g-r- C

CU CO <^ R3 E
c jc jc -T- cu
-i- oi-t-> 7 > *-
E 3 S CU 3
S- O OH-1 T3 CO
CO i- C rO
-t-> jc -i- co -a co
eu +J co s- s. E
TD 3 CU as
-o jc -o eu
co cu co S c: +->
fO t/} *r~ nj fO
3 co co O -M (j
CU >>OO CO -r-
—J O r— r—
t~*i O as X to <">
s: s- c cu
O- n3 05 "O S-
CU 05 C
JC S. Q d as C
r— CU O - CU
+-> UJ Ti— >
f^ "^x^ ^ nfl Q)
• 3 c_> +-> > co
+J CJ3 S-
-r- +j u cu eu
E CT3 •!-> JC
•r- CO C — 1 C +->
r— Ol TJ Q •!—
TJ y c^_
cai- cu o
O S- Q. U
•r- 3 • C C
+-> CO C • — . CU O
o eu 10 c ~o -i-
cu i- cu e •-- -i->
+-> tl- i — 3 4- as
CU 1 O r— C •<-
•o o o o >
•r- eu > +-> -*:
JC Ol-r- V- C T3 C
+-> S- S- fO CO S- as
CO O +J r— (J (O r—
E S- r— i. -O Jd
V4_ OJ -r- Q) C
eu o o a. a. « T3
JC ITS •*-> O
•+-> CO i — O Ol CO JC
+•) *r~ C7^ 4^
co O co as CU CU
-r- 3 -r- O «5 JC E
O" S- T- +->
_l -r- 0 r— S- S-
Q r— i — -i- O CO O
y aj u_ co M- -r- t»-

















































































•
CO
^~
-Q
TJ
o
•»—
f>»
Q.
o.
TJ

4->
o
z
 C CT>
 O .-i

 CO S-
•!- CU
 > JD
 cu E
cc: cu
    +->
    CL
    cu
    oo
IO
o
CO

-------
                             TABLE 2.
     ESTIMATED QUANTITATION LIMITS (EQL)  FOR VARIOUS MATRICES8
Matrix                                                Factor
Groundwater                                                10
Low-concentration soil by ultrasonic extraction           670
  with GPC cleanup
High-concentration soil and sludges by ultrasonic      10,000
  extraction
Non-water miscible waste                              100,000
EQL = [Method detection limit (see Table 1)] X [Factor found in this
table].   For non-aqueous samples, the factor is on a wet-weight basis.
Sample EQLs are highly matrix-dependent.  The EQLs determined herein are
provided for guidance and may not always be achievable.
                            8061  -  15                         Revision 0
                                                          September 1994

-------
                                   TABLE 3.
    AVERAGE  RECOVERIES  OF  METHOD  8061  COMPOUNDS  USING METHODS 3610 AND 3620
Compound
Dimethyl phthalate
Diethyl phthalate
Diisobutyl phthalate
Di-n-butyl phthalate
Bis(4-methyl-2-pentyl) phthalate
Bis(2-methoxyethyl) phthalate
Diamyl phthalate
Bis(2-ethoxyethyl) phthalate
Hexyl 2-ethylhexyl phthalate
Dihexyl phthalate
Benzyl butyl phthalate
Bis(2-n-butoxyethyl) phthalate
Bis(2-ethylhexyl) phthalate
Dicyclohexyl phthalate
Di-n-octyl phthalate
Dinonyl phthalate
Alumina
column8
64.5
62.5
77. 0
76.5
89.5
70.5
75.0
67.0
90.5
73.0
87.0
62.5
91.0
84.5
108
71.0
Florisil
column8
40.0
57.0
80.0
85.0
84.5
0
81.5
0
105
74.5
90.0
0
82.0
83.5
115
72.5
Alumina
cartridge13
101
103
104
108
103
64.1°
103
111
101
108
103
108
97.6
97.5
112
97.3
Florisil
cartridge*1
89.4
97.3
91.8
102
105
78. 3e
94.5
93.6
96.0
96.8
98.6
91.5
97.5
90.5
97.1
105
8 2 determinations; alumina and Florisil chromatography performed according
  to Methods 3610 and 3620, respectively.

b 2 determinations, using 1 g alumina cartridges; Fraction 1 was eluted with
  5 ml of 20-percent acetone in hexane.  40 ^9 of each component was spiked
  per cartridge.

c 36.8 percent was recovered by elution with an additional 5 ml of
  20-percent acetone in hexane.

d 2 determinations, using 1 g Florisil cartridges; Fraction 1 was eluted
  with 5 mL of 10-percent acetone in hexane.  40 jug of each component was
  spiked per cartridge.

e 14.4 percent was recovered by elution with an additional 5 ml of
  10-percent acetone in hexane.
                                   8061  -  16
    Revision 0
September 1994

-------
              c
              o
              re
              s- •
              00
             <_> 10

              o>
              a.
             oo
   <£>
   O
   00

   a
   o
                        a>
                       •(->
                        re

                       •a
                        c
                        3
                        O
                        s_
                       C3
          O)
         +->
          rO
         J=
          O
          aj
          O)
          S_
          0)
         -(->
          re
          •3.
   in
   CO

   a
   o
LU Ct:
—I O
00 U_
   
      •r-  re
       s-  s
       re TO
       3  C
      +J  3
       to  o
      LU  S-
                        0)
          O
          cO
          a;
                        CU
                       +->
                        re
                    01
                       3
                       O
                       a.
                   in in CO I — CTllOOOCTl'— i
                   cninoo<— o*3-'— icn^i"'— i

                   oincor-.r--iooj^-ooocooj'— ii— 100
i-H i-H »*«^i-H f-H i — I r— ( i — t r— 1 CSJ OJ I — I »— I I— I I— I i-H



    LO           O

ojoooioi--cr»cvicrii — OOCXI>COOCDCr>COOOCOCOCT>
                                 i— i co i— ii — i — «d-inooio«3i— i
                                 CTiCTiiocvJOO'— i i— icvir>»>— lOioinrooo
                                 •— ii— i^-icji— lexjcMevicsjcvjcvi-— ii— ii— ir— i •
                            i        ooocvitnro

                             c\iLOCMoooOf^-i~--
                             O O O"» C^ 00 00 f^
                   oo cvj un i — r-i^-
                   O"» 00 O^ CT> 00 00
                                                                                    00
                                                              1^ U> Lf>
                                                              000
                                                                                 oj .—i in
                                                           t—I •—i CO
                                                             *   •   *
                                                           in r-. co
                                                                                                  LO in
                                                                                              o «*• i—i
                                                                                              i—i o •—i
                                                              ro co oo

                                                              co co co
                                                                                              o  CTI 10
                                                                                              '-i  O O
                                                                                 en r~. o
                                                                                              CO  OJ OJ
                                                                                              IO CTl


                                                                                              OJ .—I
                                                                                              in  oo (T>

                                                                                              oo  in co
                                 CU

                                 as
                                r—
                                 
                                 rO
       O)
       -t->
       ro
       i—
       re
                                              Q.JT
                                i—  a.
                             O) >>
                             -l-> •(-> • — s
                                                        CD CD
                                                       4-> 4->
                                                        rO ro

                                                        (O ro
                                                       ^: -c
                                                        Q. Q.
                                 CU

                                 ro
                                 i— O)
                                 rC -M
                                 -C ro
                                 +J i—
                                 .C re

                             +J    +-> -I-J
                              rO -—--C  rO  CU
                             i— r— Q-i— +J
                              rO >,     ro  ro
i— ^_> +J OJ
 ro _c ^:  i
J=  Q. O-r—
+->
JZ-U
 Q--C
    a.
, —
 >-,
-C
•4->
 a) -t->
 Ea)
•i- -^
QQ
                                              a>
                                           3E
                           i  c •
                              i(/)
                            -^ -r-
                            QCQ
                    O)
                -M-t->
                 0) r0
                 ^t—
                 X re
                                                        >>X-(->-(-'-t->i — +-> rO-f->
                                          Q) i
                                          >,
                                                              ro
                                                                     X  0)
                                                                     O -C
                                                                                  ro
                                                        QJOJ
                                                       COJOOi—
                                                >,i —   i   i  i —  O  >»
                                             i — X  >>OJ OJ U  I   C
                                             >•> aj  N > — ^- — • >i c  o
                                                 CQQ
                                                       •,-  CD -r- O) -r- -I-  -1- -i- -i-
                                                       CQJCQCQCQCQQQQ
                                                                                        rO
                                                                                        CD
                                                                                        O
                                                                                       OO
    a>
   -I-J
    (O
 d) r—
+->  re
 re jc
I — •*->
 rO jz
-C  Q.
+->  o
                                                                                               Q--r- Q.
                                                           C C N
                                                           OJ QJ C
                                                          -E .C O)
                                                           Q. CLJD

                                                          Q Q Q
                                                                               O

                                                                              4->
                                                                               re
                                                                                                               0)
                                                                                                               -o
                 s-
                 re
                 •o

                 re
                 -I-J
                 
                 >


                 rO

                 "oi
                 S-
                                                                               a>
                                                                               (J
                                                                                                 a>
                                                                                                 a.
                                                                                                               01
                                                                            O)

                                                                            re
                                                                               l/l
                                                                               O)
                                                                                                               £
                                                                                                               O)
                                                                               rO
                                                                               a.
                                                                                                 OJ
                                                                                                               O)
                                                                                                 CO

                                                                                                 (/)
                                                                                                 rO
•r-  CU

 S-  i.
 O>  0)
+->  >
 O)  o
"O  U)
    CU
4-  i.
 O
    CU
 S-  O1
 O)  (O
^^f  ^_

 E  cu
 3  >
 C  
                                                                                            O i-H

                                                                                            00  i.
                                                                                           •i-  0)
                                                                                            > -Q
                                                                                            a>  E
                                                                                           a:  cu
                                                                                              4->
                                                                                               a.
                                                                                               0)
                                                                                              oo
                                                                                                                 O
                                                                                                                 CO

-------
in









(0
VO
o
00
Q
O
z
1—
UJ
2:
Q
«=C
O
LO
in
ro
o
O
:x
l—
UJ
2:
O
1 1

Q
DECISION

a
QZ




c
0
£^7
c
c "~"
o
cu
CL
oo






c
o
*- \
1 i
c d
o
o
cu
*r—
d.
OO




(«

re
oo
rO
CL
c
^
2:
cu
••-
3
UJ

O
re
oo
re
CL
o
c
3

cu
Estuarir



o


0)
^^
3


C
"^3
CU
oo

O
01


CU
Dl
3
O1


Compound sedimeni

0
CM
CM
°
in
—i
CO
S
VO
en

f— i


o
LO
ro
i — i
in
00
cu
4J
re
i—
rO
J=
j=
cu
Q
•— i •— i in oo in m ro r-~ oo «si- co
in o in •— i in co ^ CM ro ro CM
C} C^J O"^ O^ *~^ t*** ^& Lf} CO ^"^ *^
^^ ^^ t"r'3 pO CD ^VJ *sj* 00 ^^ f^* ^J * O
ovooo-H^r^cMvoi-ro^-'*
o«eMnvoinrocni^*3-ovovo
ooo en mvoroooooo
i— t * — ( i — i
inocriincoocorooO«3--^ro
CMVDCO^^-^CM^HVOCOLOCO
CMOOVO inror^.i — m en ro in
o <•<•<• ot i — •— i CM in CM oo en
i — i
CM r*^ r*. . en oo ^~ i—* ro
vororo Ln CMCMLO in
f*** oo vo o^ I"**"* o*^ LO csj
^^~ ^^ ^vj ^^ r**"* r**> vo o*^
li"> I"-*- r*«» o ^^ OO o H") t^* *-O o O*i
^Htor^^" ro •— < o> ID r*- oo vo vo
2.£.!^I^S^S2 2^2 2
vo roroi — ^CTivo ^l-OOOvo
00 lO VO r— - CM r~^ tfj> ^^ i,^ CM *5^ VO
vo ^^ co rj^ r*^ oo *o I~H cj^ co f*. ^^
• — i i — i
r-- •— i oo *a- oo CM vo oo •— i
• 	 • 	 CM LO CM •— I • 	 •— I CM
^- vo o
oo co •— i oo vo in rorO'3-
VOOCMOCMCTi O •— l i — 1
i-H l— 1 ?-H u o l—* I—* *— ' O
QJ
re QJ
i— CU +->
rO +•> QJ CU ro
•4-^ r~^ fO (fl fT3 +J
JT re i — i — -C ro
O-J= ro ro +J r—
+j -c jc x: re
• — -x: +-> ^J cu a__c
i— Q. JC .f 4-> -(->
OJQJ>i CLQ. ro— NJZ
re rO C i — • — »i — re >-,
oji — i — QJ>, i — »>QJ.C;.C: . — s
+j re re CL^: QJ >, x +->+-> +J r—
rOJZJZ i -t-i-l-JXZCUrexrcU^,
i — ^J4->CM CU rO+JXIi — CL>,X
re J= JZ i >>i— QJ i— rO X CU
-C CL CLr— X ro >,>,x:i— OJ=
+-> >>O-CXJ=-t->>)-(->r—
Sir— r— X: J= -»J O4->-C-l-> 3 >,
Q.>)>i-.^>-^J=^: Q) Cu3X>J=
4^ 4-) QJ QJ CL4_> i Xl 1 4->
i— 3 3 E E QJCM.— CO)
X! O 1 >* CM >,CM r- X >>CM CM
4J 00 C- — ~— •£•—•>> CU N »-^* —
OJ-i— c oioorooixx:coooo
QQQOQOQQCQniQCQCQCQ
t— l CM CM
CTl CTl CO
CM
t—4 i — I
Ift ^^ to
o^ r*^ oo
in *i-
CO 00 VO
if) QQ ^^
—i «*
vo vo
en oo
ro vo
en ro co
ro >—••— i
cncovo
I-- VO i— l
i — 1 r-H ^~
oo ro o
vocnoo
00
co
VO
ro
o o
CL)
rO QJ
rO ro
x: c— cu
+-> ro 4->
x: _e re
Q-4-> r—
J= ro
r— CL.C:
>> +*
X r- x:
CO >, Q.
O U i—
r— 0 >,
O i C
>, C 0
U i C
O Q Q
OJ
•^
O
01
O
rO
CU

73
ro
C
re
•M
oo
OJ

ro
CU
C
cu

QJ
S-
rO
01
renthese
re
Q

c
CD
00
CD
3
're
>
QJ
1—
ro
01
ro
^
01
C
O
+J
re
c
E
Q)
4->
CU
T3
O
CU
XI
E
3
C
CU
»—












CL
3
C
ro
CU
U
QJ
CD
cartrid
•—
to
^_
O
0
4->
QJ
O
QJ
average recoveries. All samples were sub;






e-
4->
S
o
VO
ro
-o

cu
Q.
C
re
Q)
'o
3
4-
3
I/)
O
-o
cu
CO
-O
3
00
re

o
-4->
o
re
i-
u_
' The estuarine sediment extract (Florisil,





















tetrabutyl ammonium sulfite reagent).



















rferant
o>
-^
Not able to determine because of matrix in

                                                                                                                                       O  «d-
                                                                                                                                           CM
                                                                                                                                        c  cn
                                                                                                                                        o  ^

                                                                                                                                        oo  S-
                                                                                                                                       •i-  Q)
                                                                                                                                        >  x>
                                                                                                                                        OJ  E
                                                                                                                                       or  cu
                                                                                                                                           +->
                                                                                                                                           a.
                                                                                                                                           QJ
                                                                                                                                           00
                                                                                                                                       CO
                                                                                                                                       VO
                                                                                                                                       o
                                                                                                                                       CO

-------
                                     Figure 1
                                                                DB-S
                                                                30 m x 0.53 mm 10
                                                                1.5-jimRm
        1
     6
     UJ
                          IS
                       IS
            11   12 SU-1  SU-2 SU-3
                                                    14
                                                                     16
                     SU-2 SU-3

             12  SU-1 15 T T 16
              13
                                                                 DB-1701
                                                                 30 m x 0.53 mm ID
                                                                 1.0-jim Film
                                        10
                                           11
r

^
LI
uu
L_
JL
JL
NkUJ
i i
                     10
    20

TIME (min)
                                                       30
40
GC/ECD chromatograms of a composite  phthalate esters  standard (concentration
10 ng/jiL  per compound) analyzed  on a DB-5 and a DB-1701  fused-silica open
tubular column.   Temperature program:  150°C  (0.5 min hold) to  220°C  at
5°C/min,  then to  275°C  (13 min  hold)  at 3°C/nrin.
                                    8061 -  19
                                      Revision 0
                                  September 1994

-------
                                 METHOD  8061
    PHTHALATE  ESTERS  BY  CAPILLARY  GAS  CHROMATOGRAPHY
          WITH  ELECTRON CAPTURE  DETECTION   (GC/ECD)
            Start
        7 1 Extraction
7 1 1 Refer to Chapter 2 for
     guidance on choosing
     an extraction procedure
     Recommendations given
7 1 2 Determine spike sample
     recovery and detection limit
     for each new sample matrix
     and a given extraction
     procedure.
7 1 3 Aqueous sample extraction
     with C18 disks
      1 Precondition disks using
       solvent tram
      2 Concentrate sample
       analytes on disk.
      3 Elute sample analytes
       with acetonitrile.
     4 Concentrate extract'
        1 Micro-Snyder Column
         Technique
        2 Nitrogen Slowdown
         Technique
         1 Evaporate solvent to
           desired level
         2 Rinse tube walls
           frequently and avoid
           evaporating to dryness
 7 2 Solvent Exchange to Hexane  [
 7 2 1 Evaporate extract volume to
 1 -2 ml using K-D assembly
722 Add hexane to K-D assembly
and evaporate to 1 -2 ml
 723 Rinse K-D components and
 adjust volume to desired level
                                                                        7 3 Cleanup/Fracdonation
 73.1 Cleanup may not be
      necessary for extracts with
      clean sample matrices
      Fraction collection and
      methods outlined for other
      compd groups of interest.
732 Flonsil Cartridge Cleanup
     1 Check each lot of Flonsil
       cartridges for analyte
       recovery by eluOng and
       analyzing a composite std
     2 Wash and adjust solvent
       Mow through cartridges
     3 Place culture tubes or 5 ml
       vol flasks for eluate
       collection
     4 Transfer appropriate extract
       volume on cartridge
     5 Elute the cartridges and
       dilute to mark on flask
       Transfer eluate to glass
       vials for concentration
733 Collect 2 fractions if PCBs
     and organochlonne pesticides
     are known to be present
                                                                        7 4 Gas Chromatograph
                                                                    7 4 1 Set GC operating parameters
                                                                    742 Table 1 and Figure 1 show
                                                                         MDLs and analyte retention
                                                                         times
                                   8061  -  20
                    Revision  0
              September  1994

-------
    METHOD  8061
    (CONTINUED)
       7 5 Calibration
1

75.1 See Method 8000 tor
calibration technique
7.5 2 Refer to Method 8000 tor
     internal/external std
     procedure.
       ' 6 GC Analysis
 7 6 1 Refer to Method 8000
762 Follow Section 7 6 in
     Method 8000 tor
     instructions on analysis
     sequence, dilutions.
     retention time windows.
     and identification criteria.
7 6.3  Record injection volume
      and sample peak areas
764 Identify and quantify each
     component peak using the
     internal or external std
     procedure
765 Dilute extracts which
show analyte levels
outside of the calibration
range.

1
766 Identify compounds in the
sample by comparing
retention times in the
sample and the standard
chromatograms
\
I
      8061  -  21
      Revision 0
September  1994

-------
                                 METHOD 8070

                      NITROSAMINES  BY GAS  CHROMATOGRAPHY


1.0   SCOPE AND APPLICATION

      1.1   This method covers the  determination of certain nitrosamines.   The
following compounds can be determined by this  method:


                                        	Appropriate Technique	
Compound Name               CAS No.a    3510    3520    3540    3550    3580
N-Nitrosodimethylamine 62-75-9 X X X X
N-Nitrosodiphenylamine 86-30-6 X X X X
N-Nitrosodi-n-propylamine 621-64-7 X X X X
a Chemical Abstract Services Registry Number.
X Greater than 70 percent recovery by this preparation technique.
X
X
X


      1.2   This  is  a  gas  chromatographic  (GC)  method  applicable  to  the
determination  of the  parameters  listed  above  in municipal  and  industrial
discharges.  When this method is used to analyze unfamiliar samples for any or
all of the compounds above, compound identifications  should be supported by at
least one additional qualitative technique.   This  method  describes analytical
conditions for a second gas chromatographic column  that can be used to confirm
measurements  made  with  the  primary  column.     Method   8270  provides  gas
chromatograph/mass  spectrometer   (GC/MS)   conditions  appropriate   for  the
qualitative    and    quantitative    confirmation     of    results     for
N-nitrosodi-n-propylamine.      In    order   to   confirm   the   presence   of
N-nitrosodiphenylamine,  the cleanup procedure specified  in Section  7.3.3 or 7.3.4
must be used.   In  order to confirm the presence of N-nitrosodimethylamine by
GC/MS, chromatographic column 1 of this method must be substituted  for the column
recommended  in  Method  8270.   Confirmation  of these  parameters  using  GC-high
resolution mass spectrometry or a  Thermal  Energy Analyzer  is  also  recommended
practice.

      1.3   The method detection limit  (MDL)  for each parameter is  listed in
Table 1.   The MDL  for  a  specific  wastewater may differ  from  those  listed,
depending upon the nature  of  interferences  in the sample matrix.  Table 2 lists
the Estimated Quantitation Limits (EQLs) for various matrices.

      1.4   The toxicity or carcinogenicity of each reagent used in this method
has not  been precisely defined.   However, each  chemical  compound  should  be
treated as a potential health hazard.   From this viewpoint, exposure to these
chemicals must be reduced to the lowest possible  concentration  by whatever means
available.  The  laboratory is responsible  for maintaining  a current  awareness
file of OSHA regulations regarding  the safe  handling of the chemicals  specified
in this method.  A reference  file of material  data  handling sheets  should also

                                   8070  -  1                         Revision 0
                                                                     July 1992

-------
be made available to all personnel involved in the chemical  analysis.

      1.5   These nitrosamines are known carcinogens.   Therefore,  utmost care
must be exercised  in  the handling of these materials.   Nitrosamine reference
standards and standard solutions  should  be  handled and prepared in a ventilated
glove box within a properly ventilated room.

      1.6   N-Nitrosodiphenylamine  is  reported  to  undergo  transnitrosation
reactions.  Care must  be exercised in  the heating or concentrating of solutions
containing this compound in the presence of reactive amines.


2.0   SUMMARY OF METHOD

      2.1   A measured  volume  of aqueous sample, approximately  one liter,  is
solvent extracted  with  methylene chloride using  a  separatory  funnel.   The
methylene chloride  extract is  washed with  dilute  HC1 to remove  free amines,
dried,  and  concentrated to a  volume  of 10 ml  or less.  Gas  chromatographic
conditions  are  described which permit  the separation and  measurement  of the
compounds in the extract after it has been exchanged to methanol.

      2.2   Method 8070 provides gas chromatographic conditions for the detection
of ppb  concentrations  of nitrosamines.  Prior to use  of this method, appropriate
sample  extraction techniques must be used.   Both neat and  diluted organic liquids
(Method 3580, Waste Dilution) may be analyzed  by direct  injection.  A 2 to 5 /xL
aliquot of  the extract  is injected   into  a gas  chromatograph (GC)  using the
solvent flush  technique, and  compounds in  the  GC effluent are  detected by a
nitrogen-phosphorus detector (NPD) or a Thermal Energy Analyzer and the reductive
Hall detector.
3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000.

      3.2   Matrix  interferences  may  be   caused  by  contaminants  that  are
coextracted  from the sample.   The extent  of matrix  interferences  will  vary
considerably from source to source, depending upon the nature and diversity of
the industrial  complex  or  municipality  being sampled.   The cleanup procedures
(Methods 3610 or 3620) can be used to  overcome many  of  these interferences, but
unique  samples  may require additional  cleanup  approaches to  achieve  the MDL
listed  in Table  1.

      3.3   Nitrosamines contaminate many types of products commonly found in the
laboratory.    The  analyst  must  demonstrate  that  no  nitrosamine  residues
contaminate  the sample or  solvent  extract  under the  conditions  of analysis.
Plastics, in particular, must be avoided because nitrosamines are commonly used
as  plasticizers and  are easily  extracted  from  plastic materials.   Serious
nitrosamine contamination may result at any time  if consistent quality control
is not  practiced.

      3.4   The sensitive and selective Thermal Energy Analyzer and the reductive
Hall  detector  may be used  in  place of the  nitrogen-phosphorus  detector when

                                   8070 - 2                         Revision 0
                                                                     July 1992

-------
interferences are encountered.  The Thermal Energy Analyzer offers the highest
selectivity of the non-mass spectrometric detectors.

      3.5   Solvents, reagents, glassware,  and other sample processing hardware
may yield discrete artifacts and/or elevated baselines causing misinterpretation
of gas chromatograms.  All  these materials  must be demonstrated to be free from
interferences, under the conditions of the analysis, by analyzing reagent blanks.
Specific selection of reagents and purification of solvents by distillation in
all-glass systems may be required.

      3.6   Interferences coextracted from samples will vary considerably from
source to  source,  depending upon  the  waste being sampled.   Although general
cleanup techniques are recommended as  part of  this  method,  unique samples may
require additional  cleanup.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph  - An analytical system complete with temperature
programmable gas chromatograph  suitable for on-column injection and all required
accessories including syringes, analytical  columns, gases, detector, and strip-
chart recorder.  A data system is recommended for measuring peak areas.

            4.1.1 Column 1 - 1.8 m x 4 mm ID Pyrex glass, packed with Chromosorb
      W AW, (80/100 mesh)  coated  with  10%  Carbowax  20 M/2% KOH  or equivalent.
      This  column  was used  to develop the method  performance  statements  in
      Section 9.0.   Guidelines for the  use of alternate  column  packings are
      provided in Section 7.3.2.

            4.1.2 Column  2-1.8mx4mmID  Pyrex   glass,  packed  with
      Supelcoport (100/120 mesh) coated with 10% SP-2250,  or equivalent.

            4.1.3 Detector  -   Nitrogen-Phosphorus,  reductive  Hall or  Thermal
      Energy Analyzer. These detectors have proven effective in the analysis of
      wastewaters for the parameters listed  in the scope.  A nitrogen-phosphorus
      detector was used to develop the  method performance statements in Section
      9.0.   Guidelines  for the  use  of alternate  detectors  are provided  in
      Section 7.3.2.

      4.2   Kuderna-Danish  (K-D) apparatus

            4.2.1 Concentrator tube -  10 ml, graduated  (Kontes K-570050-1025 or
      equivalent).   Calibration must be checked at the volumes employed in the
      test.  A ground glass stopper is  used to prevent  evaporation of extracts.

            4.2.2 Evaporation   flask   -  500  ml   (Kontes   K-570001-0500  or
      equivalent).     Attach to  concentrator tube  with  springs,  clamps,  or
      equivalent.

            4.2.3 Snyder  column  - Three ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.2.4 Snyder  column  -  Two ball  micro  (Kontes  K-569001-0219  or

                                   8070 - 3                         Revision 0
                                                                     July 1992

-------
      equivalent).

            4.2.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Boiling chips - Approximately 10/40 mesh.  Heat to 400°C for
30 minutes or Soxhlet extract with methylene chloride.

      4.4   Water  bath  -  Heated,  with  concentric  ring cover,  capable  of
temperature control (± 2°C).   The bath should be used in  a hood.
top.
      4.5   Balance - Analytical, 0.0001 g.

      4.6   Vials - 10 to 15 ml, amber glass with Teflon lined screw-cap or crimp
      4.7   Volumetric  flasks,  Class A,  Appropriate  sizes with  ground glass
stoppers.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used in all  tests.  Unless
otherwise indicated, it is intended that all inorganic reagents  shall conform to
the  specifications  of  the  Committee on  Analytical  Reagents  of  the  American
Chemical Society, where such specifications are  available.  Other grades may be
used, provided it is first ascertained that the  reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water - All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Methanol, CH3OH - Pesticide quality  or equivalent.

      5.4   Isooctane,  (CH3)3CCH2CH(CH3)2 - Pesticide quality  or equivalent.

      5.5   Methylene chloride, CH2C12 - Pesticide quality  or equivalent.

      5.6   Stock standard solutions (1000 mg/L) - Stock standard solutions can
be prepared from pure standard materials  or purchased as certified solutions.

            5.6.1 Prepare  stock  standard  solutions  by  accurately  weighing
      0.1000 + 0.0010 g of pure  material.  Dissolve  the material in pesticide
      quality  methanol   and  dilute  to  volume  in  a  100 ml  volumetric flask.
      Larger volumes can be used  at the  convenience of the  analyst.  If compound
      purity  is  certified at 96% or greater,  the weight can  be used without
      correction  to  calculate  the  concentration  of  the   stock  standard.
      Commercially  prepared stock standards can  be used  at any  concentration  if
      they are certified by the manufacturer or by an independent source.

            5.6.2 Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps  or  crimp tops.   Store at 4°C and protect from light.
                                   8070  - 4                         Revision 0
                                                                     July  1992

-------
      Stock  standard  solutions  should  be  checked  frequently  for  signs  of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.6.3 Stock standard solutions must be replaced after  six months, or
      sooner if comparison with check standards indicates a problem.

      5.7   Calibration standards - A minimum of five concentrations should be
prepared through dilution  of the stock standards with isooctane.  One  of the
concentrations should  be at a concentration near, but above,  the method detection
limit.  The remaining  concentrations  should correspond  to the expected range of
concentrations found in real samples or should define the working range of the
GC.   Calibration solutions  must  be  replaced after  six  months,  or  sooner if
comparison with check standards indicates a problem.

      5.8   Internal standards (if internal  standard calibration is used) - To
use this approach, the analyst must select one or more internal standards that
are similar in analytical  behavior to  the compounds  of interest.  The analyst
must further demonstrate that  the measurement of the internal  standard  is not
affected by method or matrix interferences.   Because of  these limitations, no
internal standard can be suggested that is applicable to all samples.

            5.8.1 Prepare   calibration   standards   at   a  minimum   of  five
      concentrations for each analyte of interest, as described in Section 5.7.

            5.8.2 To each calibration standard,  add a known constant amount of
      one or more internal standards, and dilute to volume with isooctane.

            5.8.3 Analyze each calibration standard according to Section 7.0.

      5.9   Surrogate standards - The analyst  should monitor the performance of
the extraction, cleanup  (when used), and analytical  system and the effectiveness
of  the  method  in  dealing  with  each  sample  matrix  by  spiking each  sample,
standard, and reagent blank with one or two surrogates (e.g. nitrosamines that
are not expected to be in  the sample)  recommended to encompass the range of the
temperature program used in this method.  Method 3500 details instructions on the
preparation of base/neutral surrogates.  Deuterated analogs of analytes should
not be  used as surrogates  for gas  chromatographic analysis due to  coelution
problems.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material  to this chapter,  Organic  Analytes,
Section  4.1.   Extracts  must be  stored  at  4°C and analyzed within 40  days of
extraction.
                                   8070 - 5                         Revision 0
                                                                     July 1992

-------
7.0   PROCEDURE

      7.1   Extraction

            7.1.1 Refer to Chapter Two  for guidance on choosing the appropriate
      extraction  procedure.    In  general,  water  samples  are  extracted at  a
      neutral, or as  is, pH with methylene chloride, using either Method 3510 or
      3520.  Solid samples are extracted using either Method 3540 or 3550.

            7.1.2 Prior to gas chromatographic analysis, the extraction solvent
      must be exchanged to methanol.   The exchange is performed during the  K-D
      procedures  listed  in all of the extraction methods.   The  exchange  is
      performed as follows.

                  7.1.2.1     Following K-D of the methylene chloride extract to
      1 ml using the macro-Snyder  column, allow the apparatus to cool and drain
      for at least 10 minutes.

                  7.1.2.2     Momentarily remove the Snyder column, add 50 ml of
            methanol, a new boiling chip, and reattach the macro-Snyder column.
            Concentrate the extract using 1 ml of methanol  to prewet the Snyder
            column.   Place the K-D  apparatus  on the  water bath  so  that  the
            concentrator tube  is  partially immersed  in  the hot water.  Adjust
            the vertical  position  of the apparatus and the water temperature, as
            required, to complete  concentration in 5-10  minutes.  At the proper
            rate of distillation the  balls of the column will actively chatter,
            but the chambers will  not flood.  When  the apparent volume of liquid
            reaches  1 ml,  remove  the K-D apparatus and allow  it  to drain  and
            cool  for at  least  10  minutes.    The  extract will   be  handled
            differently at this point,  depending on  whether or not cleanup is
            needed.  If cleanup is  not required,  proceed to  Section  7.1.2.3.  If
            cleanup  is needed, proceed to Section 7.1.2.4.

                  7.1.2.3      If cleanup of the extract  is  not required, remove
            the Snyder column and  rinse the flask and its lower joint into the
            concentrator  tube with  1-2  ml  of methanol.    A  5 ml  syringe  is
            recommended  for  this  operation.    Adjust  the extract  volume  to
            10.0 ml.  Stopper the  concentrator tube  and store refrigerated at
            4°C if further processing will not be  performed  immediately.  If the
            extract  will  be  stored  longer  than   two days,  it  should  be
            transferred to a  vial  with  a Teflon  lined screw-cap or crimp top.
            Proceed  with gas  chromatographic analysis.

                  7.1.2.4      If cleanup of the extract  is  required, remove the
            Snyder  column and rinse  the flask and  its lower joint  into  the
            concentrator tube with a minimum amount of methylene chloride.  A 5
            ml syringe is recommended for this operation.  Add a clean boiling
            chip  to  the  concentrator tube and attach  a two ball  micro-Snyder
            column.   Prewet  the  column by adding  about  0.5 ml  of methylene
            chloride to  the  top.   Place the micro K-D  apparatus  on the water
            bath  (80°C) so that the concentrator tube is partially  immersed in
            the hot  water.   Adjust the vertical  position of the apparatus and
            the water temperature, as required, to complete concentration in 5-

                                   8070  - 6                         Revision 0
                                                                     July 1992

-------
      10 minutes.   At  the proper rate of distillation the  balls  of the
      column will actively chatter, but the chambers will not flood.  When
      the  apparent volume  of liquid  reaches  0.5 ml,  remove the  K-D
      apparatus and allow it to drain and cool for at least 10 minutes.

            7.1.2.5     Remove the micro-Snyder column and  rinse the flask
      and  its  lower joint  into  the  concentrator tube with  0.2 ml  of
      methylene chloride.   Adjust  the extract volume to 2.0 ml and proceed
      with either Method 3610, 3620,  or 3640.

      7.1.3 If   N-nitrosodiphenylamine   is   to   be  measured  by   gas
chromatography, the analyst must first use a cleanup column to eliminate
diphenylamine   interference   (Methods    3610  or   3620).      If   N-
nitrosodiphenylamine is of no interest,  the analyst may proceed directly
with gas chromatographic analysis (Section 7.3).

7.2   Cleanup

      7.2.1 Cleanup procedures may not be necessary for a relatively clean
sample matrix.   The cleanup procedure  recommended  in this method has been
used for the analysis of various clean waters and industrial effluents. If
particular  circumstances  demand the use   of  an  alternative  cleanup
procedure, the analyst  must determine  the  elution profile and demonstrate
that  the recovery  of  each compound  of  interest   is  no  less  than  85%.
Diphenylamine, if present in the  original  sample extract must be separate
from  the nitrosamines  if N-nitrosodiphenylamine is to be determined by
this method.

      7.2.2 Proceed with  either  Method  3610 or  3620, using  the  2  ml
methylene chloride extracts obtained from Section  7.1.2.5.

      7.2.3 Following cleanup, the extracts should be analyzed by GC, as
described in the previous paragraphs and in Method 8000.

7.3   Gas Chromatography

      7.3.1 N-nitrosodiphenylamine completely reacts to form diphenylamine
at the normal operating temperatures of a GC  injection port (200  to 250°C).
Thus,   N-nitrosodiphenylamine  is  chromatographed   and  detected   as
diphenylamine.  Accurate determination depends on removal of diphenylamine
that  may be present in  the original  extract prior to GC  (see  Section
7.1.3).

      7.3.2 Table  1 summarizes the recommended  operating  conditions for
the gas chromatograph.   This table includes  retention  times and MDLs that
were  obtained  under   these  conditions.     Examples  of  the  parameter
separations achieved by these columns are shown in  Figures  1  and 2.  Other
packed columns,  chromatographic  conditions,  or  detectors  may  be  used if
the requirements of Section 8.2 are met.  Capillary  (open-tubular)  columns
may  also be used  if the  relative standard deviations of responses for
replicate  injections   are  demonstrated  to  be  less  than  6% and  the
requirements of Section 8.2 are met.
                             8070 - 7                         Revision 0
                                                               July 1992

-------
      7.4   Calibration - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

            7.4.1 The  procedure  for internal  or external  calibration  may  be
      used.   Refer to  Method 8000 for a description of each of these procedures.

            7.4.2 If  cleanup  is  performed on the  samples, the  analyst should
      process a  series of  standards  through  the  cleanup  procedure  and  then
      analyze the  samples by  GC.   This will confirm elution  patterns and the
      absence of interferents from the reagents.

      7.5   Gas chromatographic analysis

            7.5.1 Refer to Method 8000.   If the internal  standard calibration
      technique is used, add 10 /il_ of internal standard to the sample prior to
      injection.

            7.5.2 Method 8000 provides instructions on  the  analysis sequence,
      appropriate  dilutions,  establishing  daily  retention  time  windows,  and
      identification criteria. Include a mid-concentration check standard after
      each group of 10 samples in the analysis sequence.

            7.5.3 Examples of GC/NPD chromatograms for nitrosamines are shown in
      Figures 1 and 2.

            7.5.4 Record the sample  volume injected and the resulting peak sizes
      (in area units or peak heights).

            7.5.5 Using either the  internal  or external  calibration procedure
      (Method 8000), determine the identity  and quantity of each analyte  peak in
      the sample chromatogram.  See Method 8000 for calculation equations.

            7.5.6 If  peak  detection and  identification  are prevented  due  to
      interferences, the hexane extract may  undergo cleanup using either Method
      3610 or 3620.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter  One  for specific  quality control  procedures.
Quality control to validate sample extraction  is covered in Method 3500 and in
the extraction method utilized.   If extract cleanup was  performed,  follow the QC
in Method 3600 and in  the specific cleanup method.

      8.2   Procedures to check the GC system operation are found in Method 8000,
Section 8.6.

            8.2.1 The quality control  (QC) reference sample concentrate  (Method
      8000, Section 8.6) should contain each analyte of interest at 20 mg/L.

            8.2.2 Table 3 indicates the calibration and QC acceptance criteria
      for this method.  Table  4 gives method accuracy and precision  as  functions

                                   8070 - 8                         Revision 0
                                                                      July 1992

-------
      of concentration  for the  analytes  of interest.   The contents  of  both
      Tables should be used to  evaluate a  laboratory's  ability  to perform and
      generate acceptable data by this method.

      8.3   Calculate surrogate  standard recovery on  all  samples,  blanks, and
spikes.   Determine if the recovery is within  limits  (limits  established  by
performing QC procedures outlined in Method 8000, Section 8.10).

            8.3.1 If recovery is not within limits,  the following is required.

            •     Check to be  sure  that  there are no  errors  in calculations,
                  surrogate  solutions  and  internal  standards.   Also,  check
                  instrument performance.

            •     Recalculate the data and/or reanalyze the extract if any of
                  the above checks reveal  a problem.

            •     Reextract and reanalyze  the sample  if  none of  the above are a
                  problem or flag the data as "estimated concentration.


9.0   METHOD PERFORMANCE

      9.1   This method has been tested for linearity of recovery from spiked
organic-free reagent water and  has  been demonstrated  to be applicable for the
concentration range from 4 x MDL to 1000 x MDL.

      9.2   In a single laboratory (Southwest Research Institute), using spiked
wastewater samples, the average  recoveries presented in Table 2 were obtained.
Each spiked sample was analyzed in triplicate on  three separate  occasions.  The
standard deviation of the percent recovery is also included in Table 2.


10.0  REFERENCES

1.    Fed. Regist. 1984, 49, 43234; October 26.

2.    "Determination of Nitrosamines in Industrial and Municipal Wastewaters";
      Report for EPA Contract 68-03-2606,  in preparation.

3.    Burgess,   E.M.;  Lavanish, J.M.  "Photochemical   Decomposition  of  N-
      nitrosamines"; Tetrahedron Letters 1964, 1221.

4.    Methods for  Chemical Analysis of Water and Wastes;  U.S. Environmental
      Protection  Agency.   Office of Research  and Development. Environmental
      Monitoring and Support Laboratory.  ORD Publication Offices of Center for
      Environmental Research Information:   Cincinnati, OH,  1979; EPA-600/4-79-
      020.

5.    "Method Detection Limit and Analytical Curve Studies EPA  Methods 606, 607,
      608"; U.S. Environmental  Protection Agency. Environmental Monitoring and
      Support Laboratory,  Cincinnati, OH, special letter report for EPA Contract
      68-03-2606.

                                   8070 -  9                         Revision 0
                                                                     July 1992

-------
                                   TABLE 1.
            CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
Analyte
  Retention Time
     (minutes)	
Column 1    Column 2
                  Method
               Detection  Limit
                  (M9/L)
N-Nitrosodimethylamine
N-Ni trosodi -n-propyl ami ne
N-Ni trosodi phenyl ami nea
4.1
12.1
12. 8b
0.88
4.2
6.4C
0.15
0.46
0.81
Column 1 conditions:
   Carrier gas (He) flow rate:
   Column temperature:
Column 2 conditions:
   Carrier gas (He) flow rate:
   Column temperature:
  40 mL/min
  Isothermal,
  indicated.
  40 mL/min
  Isothermal,
  indicated.
at  110°C,   except   as  otherwise
at  120°C,   except   as  otherwise
a  Measured as diphenylamine.
b  Determined isothermally at 220°C.
c  Determined isothermally at 210°C.
                                   TABLE 2.
                    SINGLE OPERATOR ACCURACY AND PRECISION
                              Average   Standard    Spike
                              Percent   Deviation   Range
                              Number
                                of     Matrix
Analyte
Types
N-Nitrosodimethylamine
N-Ni trosodi phenyl ami ne
N-Ni trosodi -n-propyl ami ne
Recovery % (M9/L)

32
79
61

3.7
7.1
4.1

0.8
1.2
9.0
Analyses

29
29
29

5
5
5
                                   8070 - 10

-------
                                   TABLE 3.
                            QC ACCEPTANCE CRITERIA
                                    Test      Limit       Range      Recovery
                                   Cone.      for s       for X        Range
 Analyte                           (ng/L)
 N-Nitrosodimethyl amine              20        3.4       4.6-20.0     13-109

 N-Nitrosodiphenylamine              20        6.1       2.1-24.5      D-139

 N-Nitrosodi-n-propylamine           20        5.7      11.5-26.8     45-146


s     =     Standard deviation  for  four  recovery measurements,  in  yg/L.

X     =     Average recovery  for  four recovery measurements,  in  jig/L.

D     =     Detected, result  must be greater than zero.
                                   8070 -  11                         Revision 0
                                                                     July 1992

-------
                                   TABLE 4.
         METHOD  ACCURACY  AND PRECISION  AS  FUNCTIONS OF CONCENTRATION
Analyte
N-Nitrosodimethylamine
N-Ni trosodi phenyl ami ne
N-Nitroso-n-propylamine
Accuracy, as
recovery, X'
(ng/D
0.37C+0.06
0.64C+0.52
0.96C-0.07
Single
analyst
precision,
s/ (ng/L)
0.25X-0.04
0.36X-1.53
0.15X+0.13
Overall
precision,
S' (ng/L)
0.25X+0.11
0.46X-0.47
0.21X+0.15
X'    =     Expected  recovery  for  one  or  more  measurements   of  a  sample
            containing a  concentration  of C,  in  |ig/L.

sr'    =     Expected  single analyst standard deviation  of measurements at  an
            average concentration  found of  )(, in (ig/L.

C     =     True value for  the  concentration, in ng/L.

X     =     Average  recovery found  for measurements of  samples  containing  a
            concentration of C,  in jig/L.
                                   8070 - 12                         Revision 0
                                                                      July 1992

-------
             FIGURE  1.
GAS CHROMATOGRAM OF NITROSAMINES
    Column: 10% Ctrbowt* 20M + 2%
           KOH on Chromosorb W-AW
    Tfmpertturf. 110°
    Detector: Phosphorus/Nitrogtn
              6   3   10  12  14

              on tim*. minutt*
             8070  -  13
Revision 0
 July  1992

-------
                           FIGURE 2.
GAS CHROMATOGRAM  OF N-NITROSOOIPHENYLAMINE AS OIPHENYLAMINE
               Column: 10% CerboweM 20M + 2% KOH on
                      Chromosorb W-AW
               Tempertture: 220°C.
               Detector: Phosphorus/Nitrogen
                                          I
                 2   4   6   8  tO  12  14  16 19

                        Retention time, minutet
                            8070  -  14
Revision 0
 July  1992

-------
                           METHOD 8070
           NITROSAMINES  BY  GAS CHROMATOGRAPHY
                     711 Choo«e
                      apprpnale
                      ••traction
                       procedure
                     712 Perform
                    solvent exchange
                     using methanol
  7124 Perform
micro-K-0 procedure
  uting methylene
 chloride  perform
  Method 3610 or
3620.  proceed «ith
    CC analysis
              )(••
0
                          7 1 2  3 Adjuit
                        • •tract.  volun* and
                           procavd with
                         •nalyaxa or *tor*
                          in «ppropnat«
                              mannvr
                                                   713 Parform
                                                  coluan cleanup
                                                 uitng M.thod 3610
                                                     or 3620
                                                  7 3  2 Refer to
                                                    Table 1 for

                                                     operating
                                                conditions for the
                                                       CC
7 4  Refer to Method
  8000 for proper
    calibration
    technique*
  7  S  1 Refer to
  Method 8000 for
  guidance on CC
    analysis
                                           7  S 4/7 S S Record
                                             sample volume
                                             injected and
                                             resul ting peak
                                             sue/perform
                                              appropr la t«
                                           calcula tions (refer
                                             to Method 8000)
                                                 Stop
                               8070  -  15
                                                                                    Revision  0
                                                                                      July  1992

-------
                                 METHOD 8080A

            ORGANOCHLORINE  PESTICIDES AND POLYCHLORINATED BIPHENYLS
                             BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8080  is  used  to  determine  the  concentration  of  various
organochlorine pesticides and polychlorinated biphenyls (PCBs).   The  following
compounds can be determined by this method:
      Compound Name
  CAS No.'
      Aldrin
      a-BHC
      0-BHC
      5-BHC
      7-BHC (Lindane)
      Chlordane (technical)
      4,4'-DDD
      4,4'-DDE
      4,4'-DDT
      Dieldrin
      Endosulfan I
      Endosulfan II
      Endosulfan sulfate
      Endrin
      Endrin aldehyde
      Heptachlor
      Heptachlor epoxide
      4,4'-Methoxychlor
      Toxaphene
      Aroclor-1016
      Aroclor-1221
      Aroclor-1232
      Aroclor-1242
      Aroclor-1248
      Aroclor-1254
      Aroclor-1260
  309-00-2
  319-84-6
  319-85-7
  319-86-8
   58-89-9
12789-03-6
   72-54-8
   72-55-9
   50-29-3
   60-57-1
  959-98-8
33212-65-9
 1031-07-8
   72-20-8
 7421-93-4
   76-44-8
 1024-57-3
   72-43-5
 8001-35-2
12674-11-2
 1104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
a  Chemical  Abstract Services Registry Number.

      1.2   Table  1  lists the  method detection  limit for  each  compound  in
organic-free reagent water.  Table 2 lists the estimated  quantitation limit (EQL)
for other matrices.
                                  8080A - 1
                    Revision 1
                September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   Method 8080 provides gas chromatographic conditions for the detection
of ppb concentrations of certain organochlorine pesticides and PCBs.   Prior to
the use of this method, appropriate sample extraction techniques must be used.
Both neat  and diluted organic  liquids (Method 3580,  Waste Dilution)  may  be
analyzed by  direct injection.   A 2  to  5 jitl  sample  is  injected  into  a  gas
chromatograph (GC) using the  solvent  flush technique,  and  compounds  in the GC
effluent are detected by an electron capture detector (ECD) or an electrolytic
conductivity detector (HECD).

      2.2   The sensitivity of Method  8080 usually depends on the concentration
of  interferences  rather  than on  instrumental  limitations.   If interferences
prevent detection of the  analytes,  Method 8080 may also be performed on samples
that have undergone cleanup.  Method 3620,  Florisil Column Cleanup, by itself or
followed by Method 3660,  Sulfur Cleanup, may be used to eliminate interferences
in the analysis.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600,  and 8000.

      3.2   Interferences  by phthalate esters  can  pose   a  major  problem  in
pesticide  determinations when  using   the  electron  capture  detector.    These
compounds  generally  appear in  the  chromatogram as  large  late-eluting peaks,
especially  in  the 15% and  50% fractions from  the  Florisil cleanup.   Common
flexible plastics contain varying amounts of phthalates.   These phthalates are
easily extracted  or  leached  from  such materials during laboratory  operations.
Cross contamination of clean glassware  routinely occurs when plastics are handled
during extraction  steps, especially when  solvent-wetted  surfaces are handled.
Interferences from phthalates can  best  be minimized by avoiding contact with any
plastic materials.  Exhaustive cleanup  of reagents and glassware  may be required
to  eliminate  background phthalate  contamination.     The  contamination  from
phthalate  esters  can  be  completely   eliminated  with a   microcoulometric  or
electrolytic conductivity detector.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  Chromatograph:    Analytical   system  complete  with  gas
      chromatograph  suitable   for  on-column   injections   and   all   required
      accessories,  including  detectors, column  supplies,  recorder,  gases,  and
      syringes.  A data  system  for measuring peak heights and/or peak areas is
      recommended.

            4.1.2 Columns

                  4.1.2.1     Column 1:  Supelcoport  (100/120 mesh) coated with
            1.5% SP-2250/1.95% SP-2401 packed in a 1.8 m x  4  mm  ID glass column
            or equivalent.


                                   8080A  -  2                         Revision  1
                                                                September 1994

-------
                  4.1.2.2     Column 2:,  Supelcoport  (100/120 mesh) coated with
            3% OV-1 inal.8mx4mmID glass column or equivalent.

            4.1.3 Detectors:     Electron   capture   (ECD)   or   electrolytic
      conductivity detector (HECD).

      4.2   Kuderna-Danish (K-D) apparatus:

            4.2.1 Concentrator tube:  10  mL, graduated (Kontes K-570050-1025 or
      equivalent).  A  ground-glass stopper is used  to  prevent  evaporation of
      extracts.

            4.2.2 Evaporation   flask:      500   ml   (Kontes  K-570001-500   or
      equivalent).   Attach  to  concentrator  tube  with  springs,  clamps,  or
      equivalent.

            4.2.3 Snyder  column:   Three  ball  macro (Kontes K-503000-0121 or
      equivalent).

            4.2.4 Snyder  column:    Two  ball  micro  (Kontes K-569001-0219 or
      equivalent).

            4.2.5 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Boiling chips: Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.4   Water  bath:    Heated,  with  concentric  ring  cover,  capable  of
temperature control (±5°C).   The bath should be  used in  a hood.

      4.5   Volumetric flasks,  Class A:   sizes as  appropriate with ground-glass
stoppers.

      4.6   Microsyringe:  10 nl.

      4.7   Syringe:  5 ml.

      4.8   Vials:  Glass, 2, 10, and 20 ml capacity with  Teflon-lined screw caps
or crimp tops.

      4.9   Balances:   Analytical,  0.0001 g and Top loading, 0.01 g.


5.0   REAGENTS

      5.1   Reagent grade chemicals shall be used in all  tests. Unless otherwise
indicated, it is intended that  all  reagents  shall  conform to the specifications
of the Committee on Analytical  Reagents of  the American Chemical Society, where
such specifications are available. Other grades may be used, provided it is first
ascertained that  the reagent is of sufficiently high purity to  permit its use
without lessening the accuracy of the determination.
                                   8080A  - 3                         Revision 1
                                                                September 1994

-------
      5.2   Organic-free reagent water - All references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Solvents

            5.3.1 Hexane, C6H14  - Pesticide quality or equivalent.

            5.3.2 Acetone,  CH3COCH3 -  Pesticide quality or equivalent.

            5.3.3 Toluene, C6H5CH3 -  Pesticide quality or equivalent.

            5.3.4 Isooctane, (CH3)3CCH2CH(CH3)2 - Pesticide quality or equivalent.

      5.4   Stock standard solutions:

            5.4.1 Prepare  stock  standard solutions  at  a  concentration  of
      1000 mg/L  by  dissolving   0.0100  g  of  assayed  reference  material  in
      isooctane and  diluting  to volume in a  10 ml volumetric flask.   A small
      volume of toluene  may be  necessary to  put  some pesticides  in solution.
      Larger volumes can be  used  at  the convenience of  the  analyst.   When
      compound purity is  assayed to  be 96% or  greater, the weight can be used
      without correction to calculate the concentration  of the  stock standard.
      Commercially prepared stock standards can be used at any concentration if
      they are certified by the manufacturer  or by an independent source.

            5.4.2 Transfer the stock standard solutions into vials with Teflon-
      lined screw  caps  or crimp tops.   Store  at 4°C and protect  from light.
      Stock standards should be checked frequently for signs  of degradation or
      evaporation, especially just prior to preparing calibration standards from
      them.

            5.4.3 Stock standard solutions must be replaced after one year, or
      sooner if comparison with check standards indicates a problem.

      5.5   Calibration standards:   Calibration standards at  a  minimum of five
concentrations for each parameter of interest are prepared through dilution of
the stock  standards  with  isooctane.   One  of  the  concentrations  should be at a
concentration  near,  but  above, the  method  detection limit.   The remaining
concentrations should correspond to the expected range of concentrations found
in real  samples  or  should  define  the working range  of  the GC.   Calibration
solutions must be replaced after six months, or  sooner, if comparison with check
standards  indicates  a problem.

      5.6   Internal standards  (if  internal standard calibration is used):  To
use this approach, the analyst must select one or more internal  standards that
are similar in analytical behavior  to  the compounds  of  interest.   The analyst
must further demonstrate  that the measurement  of  the  internal  standard is not
affected by method  or matrix  interferences.  Because  of  these  limitations, no
internal standard can be  suggested that is applicable to all  samples.

            5.6.1 Prepare   calibration  standards  at   a  minimum   of  five
      concentrations for  each analyte of  interest as described  in Sec. 5.5.
                                   8080A -  4                         Revision 1
                                                                September 1994

-------
            5.6.2 To each calibration standard, add a known constant amount of
      one or more internal standards, and dilute to volume with isooctane.

            5.6.3 Analyze each calibration standard according to Sec. 7.0.

      5.7   Surrogate standards:  The analyst  should monitor the performance of
the extraction,  cleanup  (when used), and analytical  system and the effectiveness
of  the  method  in  dealing with each  sample  matrix  by  spiking  each  sample,
standard,  and  organic-free  reagent  water  blank  with  pesticide  surrogates.
Because GC/ECD data are much more subject to interference than GC/MS, a secondary
surrogate is  to be  used when sample  interference  is  apparent.   Two surrogate
standards (tetrachloro-m-xylene (TCMX) and decachlorobiphenyl) are added to each
sample;  however,  only  one need  be  calculated  for  recovery.    Proceed  with
corrective action when both surrogates are out  of limits for a sample (Sec. 8.3).
Method 3500 indicates the proper procedure for preparing these surrogates.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material  to this chapter,  Organic Analytes, Sec.
4.1.  Extracts must be stored under refrigeration and analyzed within 40 days of
extraction.
7.0   PROCEDURE

      7.1   Extraction:

            7.1.1 Refer to Chapter Two for guidance on choosing the appropriate
      extraction  procedure.    In  general,  water  samples  are  extracted  at  a
      neutral, or as is, pH with methylene chloride,  using either Method 3510 or
      3520.  Solid samples are extracted using Method 3540, 3541,  or 3550.

            7.1.2 Prior to gas chromatographic  analysis,  the extraction solvent
      must be exchanged to  hexane.   The exchange is performed during  the K-D
      procedures  listed  in  all of  the extraction methods.   The  exchange is
      performed as follows.

                  7.1.2.1      Following K-D of the methylene chloride extract to
            1 mL using the macro-Snyder column, allow the apparatus to cool and
            drain for at least 10 min.

                  7.1.2.2      Increase the temperature of the hot water bath to
            about 90°C.   Momentarily  remove the  Snyder  column,  add 50  mL of
            hexane,  a new boiling chip,  and reattach  the macro-Snyder column.
            Concentrate the extract using  1  mL of hexane to prewet the Snyder
            column.    Place  the K-D apparatus  on the water bath   so  that the
            concentrator tube is partially  immersed in the hot water.   Adjust
            the vertical position of the apparatus and the water temperature, as
            required, to complete concentration in 5-10 min. At the proper rate
            of distillation the balls  of the column  will  actively  chatter, but
            the chambers will  not  flood.   When the apparent volume  of liquid
                                  8080A  -  5                         Revision 1
                                                                September 1994

-------
            reaches 1 ml,  remove  the  K-D apparatus and allow  it  to  drain and
            cool for at least 10 min.

                  7.1.2.3     Remove the Snyder column and rinse the flask and
            its lower joint  into  the concentrator  tube with  1-2 ml of hexane.
            A  5 ml  syringe  is  recommended for  this operation.   Adjust  the
            extract volume to 10.0 ml.  Stopper the concentrator tube and store
            refrigerated at  4°C,  if further  processing will  not  be  performed
            immediately.  If the extract will  be stored longer than two days, it
            should be transferred to  a vial  with a Teflon-lined  screw  cap or
            crimp top.   Proceed  with  gas chromatographic  analysis  if further
            cleanup is not required.

      7.2   Gas chromatography conditions (Recommended):

            7.2.1 Column 1:

            Carrier gas (5% methane/95% argon) flow rate:    60 mL/min
            Column temperature:                              200°C  isothermal

            When  analyzing  for  the low molecular weight  PCBs  (PCB 1221-PCB
      1248), it is advisable to  set the oven  temperature to 160°C.

            7.2.2 Column 2:

            Carrier gas (5% methane/95% argon) flow rate:    60 mL/min
            Column temperature:                              200°C  isothermal

            When  analyzing  for  the low molecular weight  PCBs  (PCB 1221-PCB
      1248), it is advisable to  set the oven  temperature to 140°C.

            7.2.3 When analyzing for most or all of the analytes in this method,
      adjust the  oven temperature and  column gas flow to provide sufficient
      resolution for accurate quantitation  of  the  analytes.  This will normally
      result in a retention time  of 10  to 12 minutes for 4,4'-DDT, depending on
      the packed column used.

      7.3   Calibration: Refer to Method 8000 for proper calibration techniques.
Use Table 1  and especially Table 2 for  guidance on  selecting the lowest point on
the calibration curve.

            7.3.1 The procedure  for internal   or  external  calibration  may be
      used.   Refer to  Method 8000  for a description of each  of these procedures.

            7.3.2 Because  of the low  concentration  of   pesticide  standards
      injected on a GC/ECD, column adsorption may be a problem when the GC has
      not been  used  for a  day.   Therefore, the GC column  should  be primed or
      deactivated by injecting a PCB or pesticide standard mixture approximately
      20 times  more concentrated  than  the  mid-concentration  standard.  Inject
      this prior to beginning initial  or daily calibration.
                                   8080A -  6                         Revision 1
                                                                September 1994

-------
7.4   Gas chromatographic analysis:

      7.4.1 Refer to Method 8000.   If  the  internal  standard calibration
technique is used, add 10 /xL of internal standard to the sample prior to
injection.

      7.4.2 Method 8000 provides  instructions  on  the analysis sequence,
appropriate dilutions,  establishing daily  retention time  windows,  and
identification criteria.   Include  a mid-concentration standard after each
group of 10 samples in the analysis sequence.

      NOTE: A 72 hour sequence is not required with this method.

      7.4.3 Examples of GC/ECD chromatograms  for  various pesticides and
PCBs are shown in Figures 1 through 5.

      7.4.4 Prime the column as per Sec. 7.3.2.

      7.4.5 DDT and endrin are easily  degraded in the  injection port if
the injection port or  front of the column is dirty.  This  is  the result of
buildup  of high  boiling  residue  from sample  injection.    Check  for
degradation problems by injecting  a mid-concentration standard containing
only 4,4'-DDT and endrin.  Look for the degradation products of 4,4'-DDT
(4,4'-DDE and 4,4'-ODD) and endrin (endrin  ketone and  endrin aldehyde).
If degradation of  either DDT or endrin exceeds 20%, take corrective action
before proceeding  with  calibration, by following the  GC system maintenance
outlined in of Method 8000.  Calculate percent breakdown as follows:

                          Total DDT degradation peak area (DDE + ODD)
      % breakdown       =	  x 100
      for 4,4'-DDT          Total DDT peak area (DDT + DDE + ODD)

                          Total endrin degradation peak area
                          (endrin aldehyde + endrin ketone)
      % breakdown       = 	   x 100
      for Endrin           Total  endrin peak area (endrin +
                           endrin aldehyde + endrin ketone)

      7.4.6 Record the sample volume injected and the resulting peak sizes
(in area units or peak heights).

      7.4.7 Using either  the  internal  or  external calibration procedure
(Method 8000), determine the  identity and quantity of each component peak
in the  sample chromatogram which corresponds  to  the compounds  used for
calibration purposes.

      7.4.8 If peak  detection and  identification are  prevented  due to
interferences, the hexane  extract may need to undergo cleanup using Method
3620.   The  resultant  extract(s)  may be analyzed  by GC  directly  or may
undergo further cleanup to remove sulfur using Method 3660.
                             8080A  -  7                         Revision 1
                                                          September 1994

-------
                                          '0
7.5   Cleanup:

      7.5.1 Proceed with Method 3620,  followed  by,  if necessary,  Method
3660, using the 10 ml hexane extracts obtained from Sec. 7.1.2.3.

      7.5.2 Following cleanup, the extracts should be analyzed by GC, as
described in the previous sections and in Method 8000.

      7.5.3 If only  PCBs are to  be  measured  in a  sample,  the sulfuric
acid/permanganate  cleanup   (Method  3665),  followed  by Silica  Cleanup
(Method 3630) or Florisil Cleanup (Method 3620), is recommended.

7.6   Calculations (excerpted from U.S. FDA, RAM):

      7.6.1 Calculation of Certain Residues:  Residues which are mixtures
of two or more components present  problems  in measurement.  When they are
found together,  e.g., toxaphene  and DDT,  the  problem  of  quantitation
becomes even more difficult.   In  the following  sections suggestions are
offered for  handling toxaphene,  chlordane,  PCB,  DDT,  and  BHC.    A 10%
DC-200 stationary phase  column  was used to obtain  the  chromatograms in
Figures 6-9.

      7.6.2 Toxaphene:  Quantitative  calculation of toxaphene or Strobane
is difficult,  but reasonable  accuracy can be  obtained.   To  calculate
toxaphene on GC/ECD:   (a) adjust sample size so that toxaphene major peaks
are  10-30% full-scale deflection  (FSD);  (b)  inject  a toxaphene standard
that is estimated to  be  within +10 ng of  the sample;  (c)  construct the
baseline of standard toxaphene between its  extremities;  and  (d) construct
the baseline under the sample, using  the  distances of the peak troughs to
baseline on the standard  as  a guide (Figures 7, 8,  and  9).  This procedure
is made difficult by the  fact  that the relative  heights  and widths of the
peaks in  the  sample  will probably not be  identical to  the  standard.  A
toxaphene standard  that  has been passed through  a  Florisil  column will
show a shorter retention time for peak X and an  enlargement of peak Y.

      7.6.3 Toxaphene and  DDT:   If DDT  is  present,  it  will superimpose
itself on toxaphene peak V.   To  determine the approximate baseline of the
DDT, draw a line  connecting  the  trough of peaks U and V with  the trough of
peaks W and X and construct another line  parallel to this line which will
just cut  the  top of  peak W  (Figure 61).   This procedure was tested with
ratios of standard toxaphene-DDT mixtures from 1:10 to 2:1 and the results
of added  and calculated  DDT and toxaphene by the "parallel  lines" method
of baseline  construction were  within 10% of the actual values  in all
cases.

            7.6.3.1      A   series  of  toxaphene   residues   have  been
      calculated using total peak area for  comparison to the standard and
      also  using  area of  the last four peaks  only in  both  sample and
      standard.   The  agreement between the results  obtained  by  the two
      methods  justifies  the  use  of  the  latter method  for calculating
      toxaphene  in   a  sample  where   the  early  eluting  portion   of the
      toxaphene chromatogram  is interfered with by other substances.
8080A - 8                         Revision  1
                              September  1994

-------
            7.6.3.2     The  baseline  for  methoxychlor superimposed  on
      toxaphene (Figure 8b) was constructed by overlaying the samples on
      a toxaphene  standard of approximately the same concentration (Figure
      8a) and viewing the charts against a lighted background.

      7.6.4 Chlordane  is  a  technical  mixture  of  at least  11  major
components  and  30   or  more   minor   ones.     Gas  chromatography-mass
spectrometry and  nuclear magnetic resonance  analytical  techniques have
been applied to the  elucidation of the chemical  structures  of the many
chlordane constituents.  Figure 9a  is  a chromatogram of standard chlor-
dane.  Peaks E and F  are responses  to  trans-  and cis-chlordane,  respec-
tively.  These are the two major  components  of technical  chlordane, but
the exact percentage of each in the technical  material is not completely
defined and is not consistent from batch to batch.  Other labelled peaks
in  Figure  9a are  thought  to represent:   A,  monochlorinated  adduct  of
pentachlorocyclopentadiene   with   cyclopentadiene;   B,   coelution   of
heptachlor and a-chlordene; C,  coelution of /3-chlordene and 7-chlordene;
D, a chlordane analog; G, coelution of cis-nonachlor and "Compound K," a
chlordane isomer.  The right "shoulder" of peak  F is   caused by trans-
nonachlor.

            7.6.4.1     The GC pattern of a chlordane residue may differ
      considerably from that  of the  technical  standard.  Depending on the
      sample substrate and its  history,  residues  of chlordane can consist
      of  almost   any  combination  of  constituents  from  the  technical
      chlordane,   plant  and/or  animal  metabolites,  and  products  of
      degradation  caused by  exposure  to  environmental factors  such  as
      water and sunlight.  Only limited  information is  available on which
      residue GC  patterns are likely to  occur  in  which  samples types, and
      even this information may not be applicable to a situation where the
      route of exposure is  unusual.  For example,  fish exposed to a recent
      spill  of technical chlordane will  contain a  residue  drastically
      different from  a  fish  whose chlordane  residue was  accumulated by
      ingestion of smaller  fish  or of vegetation,   which  in turn  had
      accumulated  residues  because  chlordane  was  in the  water  from
      agricultural runoff.

            7.6.4.2     Because of this inability to predict a chlordane
      residue GC  pattern, it  is not  possible to prescribe a single method
      for the quantitation of chlordane  residues. The  analyst must judge
      whether or  not the residue's GC pattern is sufficiently similar to
      that of a technical chlordane reference material to use the latter
      as a reference standard for quantitation.

            7.6.4.3     When  the  chlordane  residue  does  not  resemble
      technical chlordane, but instead consists primarily of individual,
      identifiable peaks,  quantitate each  peak separately  against  the
      appropriate reference materials and  report  the  individual residues.
      (Reference   materials  are available  for  at  least  11  chlordane
      constituents, metabolites or degradation products which may occur in
      the residue.)
                             8080A  -  9                         Revision 1
                                                          September 1994

-------
            7.6.4.4     When the  GC pattern of the residue resembles that
      of technical  chlordane,  quantitate chlordane residues by comparing
      the total  area  of  the  chlordane chromatogram from peaks A through F
      (Figure 9a)  in the sample  versus the same  part  of  the  standard
      chromatogram.  Peak G may be obscured in a sample by the presence of
      other  pesticides.    If  G   is  not  obscured,   include  it  in  the
      measurement for both standard and sample.  If the heptachlor epoxide
      peak is relatively small, include  it as part of the total chlordane
      area  for   calculation  of   the  residue.    If  heptachlor  and/or
      heptachlor epoxide  are  much out  of proportion as  in  Figure  6j,
      calculate  these separately  and subtract their areas from total area
      to give a  corrected chlordane  area.   (Note  that octachlor epoxide,
      a metabolite of chlordane,  can  easily be mistaken  for  heptachlor
      epoxide on a nonpolar  GC column.)

            7.6.4.5     To  measure  the  total  area  of  the  chlordane
      chromatogram,  proceed as  in Sec.  7.6.2 on  toxaphene.   Inject  an
      amount  of  technical  chlordane   standard  which  will  produce  a
      chromatogram in which  peaks E and  F are approximately the same size
      as  those  in the  sample chromatograms.    Construct  the  baseline
      beneath the standard from the beginning of peak A to  the end of peak
      F as shown in Figure 9a.  Use  the  distance  from the  trough between
      peaks E and F to the baseline in  the  chromatogram of the standard to
      construct  the baseline in the chromatogram of the sample.  Figure 9b
      shows  how  the   presence of toxaphene  causes  the  baseline  under
      chlordane  to take  an upward angle.  When  the size  of peaks E and F
      in standard and sample chromatograms are the  same, the distance from
      the  trough of  the peaks  to  the  baselines should  be  the  same.
      Measurement of  chlordane area  should be  done by total peak area if
      possible.

            NOTE: A  comparison  has  been  made of the  total  peak area
                  integration method  and  the addition  of  peak  heights
                  method for  several samples containing chlordane.  The
                  peak heights A,  B,  C,  D,  E,  and  F  were measured  in
                  millimeters from peak maximum of  each to the baseline
                  constructed under the total chlordane area and were then
                  added   together.   These  results  obtained by  the  two
                  techniques are  too close to ignore  this method of "peak
                  height addition" as a means of  calculating  chlordane.
                  The technique has inherent difficulties  because not all
                  the peaks  are symmetrical and not all are present  in the
                  same ratio in standard and in sample.  This method does
                  offer   a means  of  calculating  results  if no  means  of
                  measuring  total area is practical.

      7.6.5 Polychlorinated  biphenyls  (PCBs):  Quantitation of  residues of
PCB involves problems similar  to  those encountered in the quantitation of
toxaphene, Strobane,  and chlordane.  In each case,  the chemical is made up
of numerous compounds.   So the chromatograms are multi-peak. Also in each
case, the chromatogram of the  residue may  not match that of the standard.

            7.6.5.1     Mixtures of PCBs of various chlorine contents were
      sold  for  many  years  in the U.S.  by the  Monsanto   Co.  under  the

                            8080A -  10                        Revision 1
                                                          September 1994

-------
      tradename Aroclor (1200 series and 1016).  Though these Aroclors are
      no  longer  marketed,  the  PCBs  remain  in  the environment  and are
      sometimes found as residues in foods, especially fish.

            7.6.5.2     PCB residues are quantitated by comparison to one
      or more of the Aroclor materials, depending on the chromatographic
      pattern of the residue.  A choice must be made as to which Aroclor
      or mixture of Aroclors will produce a chromatogram most similar to
      that of the residue.   This may  also  involve a judgment about what
      proportion of  the different  Aroclors to  combine  to  produce the
      appropriate reference material.

            7.6.5.3     Quantitate PCB residues by comparing total area or
      height  of  residue peaks  to  total  area of  height  of  peaks  from
      appropriate Aroclor(s) reference materials.  Measure total area or
      height  response  from common  baseline  under all  peaks.   Use only
      those   peaks   from  the   sample  that   can  be   attributed  to
      chlorobiphenyls.    These  peaks  must  also  be  present  in  the
      chromatogram of the reference materials.  Mixtures of Aroclors may
      be required to provide the best match of GC patterns of sample and
      reference.

      7.6.6 DDT:  DDT  found  in  samples often consists  of  both o,p'- and
p,p'-DDT.   Residues  of DDE  and  ODD  are also frequently  present.   Each
isorner of  DDT and  its metabolites  should  be  quantitated  using the pure
standard of that compound and reported as such.

      7.6.7 Hexachlorocyclohexane   (BHC,  from the  former  name, benzene
hexachloride):   Technical  grade BHC is a  cream-colored  amorphous  solid
with a very  characteristic musty odor; it consists of  a  mixture of six
chemically distinct isomers and one or more heptachloro-cyclohexanes and
octachloro-cyclohexanes.

            7.6.7.1     Commercial   BHC  preparations  may  show a  wide
      variance  in  the percentage  of  individual  isomers  present.    The
      elimination rate of the isomers  fed to  rats was 3 weeks for the a-,
      7-,  and  5-isomers  and 14 weeks  for the /3-isomer.   Thus it may be
      possible to have any  combination of the various isomers in different
      food commodities.  BHC found in dairy products usually has a large
      percentage of /3-isomer.

            7.6.7.2     Individual  isomers  (a,  0, 7, and 6) were injected
      into   gas   chromatographs  equipped   with   flame   ionization,
      microcoulometric, and electron  capture detectors.  Response for the
      four isomers is  very nearly  the same  whether flame  ionization or
      microcoulometric GLC is used.  The a-, 7-,  and 
-------
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality  control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was  performed, follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Quality control required to evaluate the GC system operation is found
in Method 8000.

            8.2.1 The quality  control  check sample concentrate  (Method  8000)
      should  contain  each  single-component   parameter  of  interest  at  the
      following  concentrations  in  acetone  or  other  water  miscible  solvent:
      4,4'-DDD, 10 mg/L; 4,4'-DDT, 10 mg/L;  endosulfan II,  10 mg/L; endosulfan
      sulfate,  10  mg/L;   endrin,   10 mg/L;  and  any  other  single-component
      pesticide, 2 mg/L.  If this method is only to be used to analyze for PCBs,
      chlordane, or  toxaphene, the  QC check sample concentrate should contain
      the most representative multi-component parameter at a concentration of 50
      mg/L in acetone.

            8.2.2 Table 3 indicates  the  QC acceptance criteria for this method.
      Table 4 gives method accuracy  and  precision  as functions of concentration
      for the analytes of interest.   The contents  of both Tables should be used
      to evaluate a laboratory's ability to perform and  generate acceptable data
      by this method.

      8.3   Calculate surrogate standard  recovery on  all  samples,  blanks,  and
spikes.   Determine  if  the  recovery is within limits (limits  established by
performing QC procedures outlined in Method 8000).

            8.3.1  If recovery is not within limits, the following is required.

                  •     Check to be  sure  there are no  errors in calculations,
                        surrogate solutions  and internal standards.  Also, check
                        instrument performance.

                  •     Recalculate the data and/or reanalyze the extract if any
                        of the above checks reveal a problem.

                  •     Reextract and reanalyze the sample  if none of the above
                        are  a   problem  or  flag  the  data   as   "estimated
                        concentration".

      8.4   GC/MS confirmation: Any compounds confirmed by  two  columns may  also
be confirmed by GC/MS if the  concentration is sufficient for detection by GC/MS
as determined by the laboratory generated detection limits.

            8.4.1 The GC/MS would  normally require a minimum concentration of 10
      ng/jLtL  in  the final  extract, for each  single-component  compound.

            8.4.2 The pesticide extract and  associated  blank  should be analyzed
      by GC/MS  as  per Sec. 7.0 of Method  8270.
                                  8080A - 12                        Revision 1
                                                                September 1994

-------
            8.4.3 The  confirmation may  be  from the  GC/MS  analysis  of  the
      base/neutral-acid extractables extracts (sample and blank).  However, if
      the  compounds  are  not  detected in  the base/neutral-acid  extract even
      though the concentration is high enough, a GC/MS analysis of the pesticide
      extract should be performed.

            8.4.4 A reference standard of the compound must also be analyzed by
      GC/MS.  The  concentration  of the reference standard must  be  at a level
      that  would  demonstrate  the ability   to  confirm  the  pesticides/PCBs
      identified by GC/ECD.


9.0   METHOD PERFORMANCE

      9.1   The method was tested by 20 laboratories  using organic-free reagent
water, drinking water,  surface water, and three industrial wastewaters  spiked at
six concentrations.  Concentrations used in the study ranged from 0.5 to 30 |ug/L
for single-component pesticides  and from 8.5 to 400 jug/L  for multi-component
parameters.  Single operator  precision, overall precision,  and method accuracy
were  found  to  be directly related  to the concentration of  the  parameter and
essentially independent of the sample matrix.   Linear equations to describe these
relationships for an electron capture detector are presented  in Table 4.

      9.2   The accuracy and  precision obtained will  be determined  by the  sample
matrix,   sample-preparation   technique,   optional   cleanup   techniques,  and
calibration procedures used.


10.0  REFERENCES

1.    U.S.  EPA,  "Development  and Application of Test Procedures  for Specific
      Organic Toxic  Substances  in Wastewaters,  Category  10: Pesticides  and
      PCBs," Report for EPA Contract 68-03-2605.

2.    U.S.  EPA,  "Interim Methods  for the Sampling  and Analysis of Priority
      Pollutants in  Sediments and Fish Tissue,"  Environmental Monitoring and
      Support Laboratory, Cincinnati,  OH 45268, October 1980.

3.    Pressley,  T.A.,  and J.E.  Longbottom,  "The Determination of Organohalide
      Pesticides and PCBs in  Industrial and Municipal Wastewater: Method 617,"
      U.S. EPA/EMSL,  Cincinnati, OH, EPA-600/4-84-006, 1982.

4.    "Determination  of  Pesticides  and  PCB's  in   Industrial  and  Municipal
      Wastewaters,   U.S.   Environmental   Protection  Agency,"   Environmental
      Monitoring and  Support Laboratory, Cincinnati,  OH 45268, EPA-600/4-82-023,
      June 1982.

5.    Goerlitz,  D.F.  and L.M.  Law,  Bulletin for Environmental Contamination and
      Toxicology, 6,  9, 1971.

6.    Burke, J.A.,  "Gas Chromatography  for  Pesticide Residue Analysis; Some
      Practical  Aspects,"  Journal  of the  Association of Official  Analytical
      Chemists,  48, 1037, 1965.


                                  8080A - 13                        Revision 1
                                                                September 1994

-------
7.    Webb,  R.G.  and A.C.  McCall,  "Quantitative  PCB Standards  for Electron
      Capture Gas Chromatography," Journal  of Chromatographic Science, JJ_, 366,
      1973.

8.    Millar, J.D.,  R.E.  Thomas  and H.J.  Schattenberg,  "EPA  Method Study 18,
      Method 608: Organochlorine Pesticides and PCBs," U.S. EPA/EMSL, Research
      Triangle Park, NC, EPA-600/4-84-061, 1984.

9.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants Under the Clean Water Act; Final  Rule and Interim
      Final Rule and Proposed Rule," October 26, 1984.

11.   U.S. Food  and  Drug  Administration,  Pesticide Analytical  Manual, Vol. 1,
      June 1979.

12.   Sawyer, L.D., JAOAC, 56, 1015-1023 (1973), 61 272-281 (1978), 61 282-291
      (1978).

13.   Stewart, J.  "EPA Verification Experiment for Validation of the SOXTEC* PCB
      Extraction  Procedure";  Oak Ridge National  Laboratory,  Oak  Ridge,  TN,
      37831-6138; October 1988.
                                   8080A  -  14                         Revision  1
                                                                 September 1994

-------
                                   TABLE 1.
                  GAS CHROMATOGRAPHY OF PESTICIDES AND PCBsa
Analyte
Aldrin
a-BHC
0-BHC
5-BHC
7-BHC (Lindane)
Chlordane (technical)
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
Methoxychlor
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
Retention
Col. 1
2.40
1.35
1.90
2.15
1.70
e
7.83
5.13
9.40
5.45
4.50
8.00
14.22
6.55
11.82
2.00
3.50
18.20
e
e
e
e
e
e
e
e
time (min)
Col. 2
4.10
1.82
1.97
2.20
2.13
e
9.08
7.15
11.75
7.23
6.20
8.28
10.70
8.10
9.30
3.35
5.00
26.60
e
e
e
e
e
e
e
e
Method
Detection
limit (Mg/L)
0.004
0.003
0.006
0.009
0.004
0.014
0.011
0.004
0.012
0.002
0.014
0.004
0.066
0.006
0.023
0.003
0.083
0.176
0.24
nd
nd
nd
0.065
nd
nd
nd
aU.S.  EPA.   Method  617.   Organochlorine  Pesticides and  PCBs.
Monitoring and Support Laboratory, Cincinnati, Ohio 45268.

e  =  Multiple peak response.

nd = not determined.
 Environmental
                                  8080A - 15
    Revision 1
September 1994

-------
                                  TABLE 2.
DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQLs) FOR VARIOUS MATRICES8
        Matrix                                                Factor
        Ground water                                               10
        Low-concentration soil  by sonication with GPC cleanup     670
        High-concentration soil  and sludges by sonication      10,000
        Non-water miscible waste                              100,000
           EQL = [Method detection limit (see Table 1)]  X  [Factor found in this
           table].   For  non-aqueous samples, the  factor is on  a wet-weight
           basis.  Sample EQLs are  highly matrix-dependent.   The EQLs listed
           herein are provided for guidance  and  may not always be achievable.
                                 8080A - 16                        Revision 1
                                                               September 1994

-------
                                   TABLE 3.
                            QC ACCEPTANCE CRITERIA8
Analyte
Aldrin
a-BHC
0-BHC
6-BHC
7-BHC
Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrin
Heptachlor
Heptachlor epoxide
Toxaphene
PCB-1016
PCB-1221
PCB-1232
PCB-1242
PCB-1248
PCB-1254
PCB-1260
Test
cone.
(M9/L)
2.0
2.0
2.0
2.0
2.0
50
10
2.0
10
2.0
2.0
10
10
10
2.0
2.0
50
50
50
50
50
50
50
50
Limit
for s
(M9/L)
0.42
0.48
0.64
0.72
0.46
10.0
2.8
0.55
3.6
0.76
0.49
6.1
2.7
3.7
0.40
0.41
12.7
10.0
24.4
17.9
12.2
15.9
13.8
10.4
Range
for x
(M9/L)
1.08-2.24
0.98-2.44
0.78-2.60
1.01-2.37
0.86-2.32
27.6-54.3
4.8-12.6
1.08-2.60
4.6-13.7
1.15-2.49
1.14-2.82
2.2-17.1
3.8-13.2
5.1-12.6
0.86-2.00
1.13-2.63
27.8-55.6
30.5-51.5
22.1-75.2
14.0-98.5
24.8-69.6
29.0-70.2
22.2-57.9
18.7-54.9
Range
P> PS
(%)
42-122
37-134
17-147
19-140
32-127
45-119
31-141
30-145
25-160
36-146
45-153
D-202
26-144
30-147
34-111
37-142
41-126
50-114
15-178
10-215
39-150
38-158
29-131
8-127
s     =     Standard deviation of four recovery measurements, in /xg/L.

x     =     Average recovery for four recovery measurements, in M9/L.

P, Ps =     Percent recovery measured.
D
Detected; result must be greater than zero.
"Criteria from 40 CFR Part 136 for Method 608.  These criteria are based directly
upon the method performance  data in  Table  4.   Where necessary, the limits for
recovery  have  been  broadened  to  assure applicability   of  the  limits  to
concentrations below those used to develop Table 4.
                                  8080A - 17
                                                        Revision 1
                                                    September 1994

-------
                                   TABLE 4.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION8
Analyte
Aldrin
a-BHC
0-BHC

-------
               Figure 1
      Gas Chromatogram of Pesticides
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
              I        12
       WTlNT1ON TIME (MINUTES)
II
              8080A - 19
            Revision 1
         September 1994

-------
            Figure 2
  Gas Chromatogram of Chlordane
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
     4        I        12

    RCTENTION TlMf (MINUTES)
16
           8080A - 20
         Revision  1
      September 1994

-------
             Figure  3
    Gas  Chromatogram of Toxaphene
              Column:    1.5% SP-2250+
                         1.95% SP-2401  on Supelcoport
              Temperature:  200°C
              Detector:  Electron Capture
f       10       14       II

     NfTtNTION TIM! (MINUTIS)
22
26
            8080A - 21
           Revision 1
       September 1994

-------
                 Figure 4
      Gas Chromatogram of Aroclor 1254
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
       •         10        14

         MITtNTlON TIMI (MINUTES)
It
22
                8080A - 22
          Revision 1
      September  1994

-------
                      Figure 5
           Gas  Chromatogram of Aroclor 1260
Column:    1.5% SP-2250+
           1.95% SP-2401  on Supelcoport
Temperature:  200°C
Detector:  Electron Capture
               10       U       II
             MITINTION TIMI (MINUTU)
                     8080A - 23
    Revision  1
September 1994

-------
                            Figure 6
           1   a
        J..L
Fig.6--Baseline construction for some typical gas chromotagraphic  peaks.
a: symmetrical separated flat baseline;  b and c:  overlapp flat  baseline;
d: separated (pen does not return to baseline between peaks); e: separated
sloping baseline; f: separated (pen goes below baseline between peaks);
g: a- and 7-BHC  sloping baseline;  h: a-,0-  and 7-BHC sloping baseline;
i: chlordane flat baseline; j: heptachlor and heptachlor epoxide super-
imposed on chlordane; k: chair-shaped peaks,  unsymmetrical peak;
1: p,p'-DDT superimposed on toxaphene.
                           8080A - 24
    Revision 1
September 1994

-------
                               Figure 7
 Fig.-  7a -- Baseline construction for multiple residues with standard
                              toxaphene.
Fig.- 7b -- Baseline construction for multiple residues with toxaphene,
                      DDE and o,p'-, and p,p'-DDT
                                8080A -  25
    Revision 1
September 1994

-------
                                  Figure 8
  Fig.- 8a -- Baseline construction for multiple residues: standard toxaphene.
Fig.- 8b --
Baseline construction for multiple residues: rice bran with BHC,
         toxaphene, DDT, and methoxychlor.
                               8080A - 26
                                                     Revision 1
                                                 September 1994

-------
                               Figure 9
Fig.- 9a -- Baseline construction for multiple residues: standard chlordane.
Fig.- 9b -- Baseline construction for multiple residues: rice bran with
                     chlordane,  toxaphene,  and DDT.
                              8080A - 27
    Revision 1
September 1994

-------
                                METHOD 8080A
       ORGANOCHLORINE  PESTICIDES AND  POLYCHLORINATED BIPHENYLS
                           BY GAS  CHROMATOGRAPHY
       Start
     7.1.1 Choose
appropriate extraction
      procedure.
 7.3.2 Prime or
deactivate the GC
 column prior to
 daily calibration.
   7.1.2 Exchange
 extraction solvent
     to hexane.
       7.2 Set
   chromatographic
      conditions.
      7.3 Refer to
   Method 8000 for
  proper calibration
      techniques.
  7.4 Perform
  GC analysis.
     7.4.8
     is peak
  detection and
   identification
   prevented?
     7.6.1 Do
  residues have
    two or more
   components?
   7.5.1 Cleanup
 using Method 3620
or 3660 if necessary.
    7.6 Calculate
   concentrations.
                                 8080A - 28
                                  Revision  1
                             September  1994

-------
                                  METHOD  8081

             ORGANOCHLORINE  PESTICIDES AND  PCBs AS AROCLORS  BY GAS
                  CHROMATOGRAPHY: CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8081  is used  to determine  the  concentrations  of  various
organochlorine pesticides and polychlorinated biphenyls (PCBs) as Aroclors, in
extracts from solid and liquid matrices.  Open-tubular, capillary columns were
employed with  electron capture detectors  (ECD) or  electrolytic conductivity
detectors (ELCD).  When compared to the packed columns,  these fused-silica, open-
tubular  columns  offer  improved  resolution,  better  selectivity,  increased
sensitivity, and faster analysis.  The list below is annotated to show whether a
single- or dual-column  analysis  system was used to  identify each target analyte.
            Compound Name
CAS Registry No.
Aldrina-b
Aroclor-1016a'b
Aroclor-1221a'b
Aroclor-1232a'b
Aroclor-1242a'b
Aroclor-1248a-b
Aroclor-1254a-b
Aroclor-1260a'b
a-BHCa'b
j8-BHCa'b
7-BHC (Lindane)a'b
5-BHCa-b
Chlorobenzilateb
a-Chlordaneb
7-Chlordanea'b
DBCPb
4,4'-DDDa-b
4,4'-DDEa'b
4,4'-DDTa'b
Diallateb
Dieldrina'b
Endosulfan Ia-b
Endosulfan IIa-b
Endosulfan sulfatea'b
Endrina'b
Endrin aldehydea'b
Endrin ketoneb
309-00-2
12674-11-2
1104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
319-84-6
319-85-7
58-89-9
319-86-8
510-15-6
5103-71-9
5103-74-2
96-12-8
72-54-8
72-55-9
50-29-3
2303-16-4
60-57-1
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
                                   8081  -  1
                          Revision 0
                      September 1994

-------
            Compound Name                 CAS Registry No.
             Heptachlora'b                     76-44-8
             Heptachlor epoxidea-b           1024-57-3
             Hexachlorobenzeneb              118-74-1
             Hexachlorocyclopentadieneb       77-47-4
             Isodrin6                        465-73-6
             Keponeb                         143-50-0
             Methoxychlora-b                   72-43-5
             Toxaphenea'b                    8001-35-2
                Single-column analysis
                Dual-column analysis
      1.2   The analyst must select columns, detectors and calibration procedures
most appropriate  for  the specific analytes  of  interest in a  study.   Matrix-
specific performance data must be established  and the stability of the analytical
system and instrument  calibration must be established for each analytical matrix
(e.g.,  hexane solutions from sample extractions, diluted oil samples, etc.).

      1.3   Although  performance  data  are  presented for  many of  the listed
chemicals,  it  is  unlikely that  all  of them  could  be determined  in  a single
analysis.   This limitation results  because  the chemical  and chromatographic
behavior  of many  of  these  chemicals  can  result  in  co-elution.    Several
cleanup/fractionation  schemes are  provided in this  method  and in Method 3600.
Any chemical is a  potential method  interference when it is not  a target analyte.

      1.4   Several multi-component mixtures  (i.e.,  Aroclors and Toxaphene) are
listed as target compounds.  When  samples  contain more than  one multi-component
analyte, a  higher  level  of  analyst expertise is required  to attain acceptable
levels of  qualitative  and  quantitative  analysis.    The  same is  true of multi-
component  analytes  that  have been subjected to environmental  degradation or
degradation by treatment  technologies.  These result in "weathered" Aroclors (or
any other multi-component mixtures) that may have significant differences in peak
patterns than those of standards.   In  these cases, individual congener analyses
may be preferred over  total mixture analyses.

      1.5   Compound  identification based on  single column analysis should be
confirmed  on  a second column,  or should be  supported  by   at  least  one other
qualitative technique. This method describes analytical  conditions  for  a second
gas chromatographic column that can be used to confirm the measurements made with
the primary column.   GC/MS Method 8270 is also  recommended as a confirmation
technique  if sensitivity permits  (Sec. 8).

      1.6   This method  describes  a dual  column option.   The option allows  a
hardware configuration of two analytical  columns  joined to a single injection
port.   The option allows one injection to  be used for dual  column analysis.

                                   8081 - 2                         Revision  0
                                                                September  1994

-------
Analysts are cautioned that the dual column option may not be appropriate when
the instrument is subject to mechanical  stress,  many  samples  are  to  be run in a
short period, or when contaminated samples are analyzed.

      1.7   This method  is restricted to  use  by or under the  supervision  of
analysts  experienced  in  the  use  of  a  gas chromatograph  (GC)  and  in  the
interpretation of gas chromatograms.   Each  analyst must demonstrate  the ability
to generate acceptable results with this method.

      1.8   Extracts suitable for analysis by this method may also  be analyzed
for  organophosphorus pesticides  (Method 8141).   Some  extracts may  also  be
suitable for triazine  herbicide  analysis,  if low recoveries  (normally samples
taken for triazine analysis must be preserved) are not a problem.

      1.9   The following compounds may also be determined using this method:
            Compound Name
CAS Registry No.
             Alachlora'b
             Captafolb
             Captanb
             Chloronebb
             Chloropropylateb
             Chlorothalonilb
             DCPAb
             Dichloneb
             Dicofolb
             Etridiazoleb
             Halowax-1000b
             Halowax-1001b
             Halowax-1013b
             Halowax-1014b
             Halowax-1051b
             Halowax-1099b
             Mirexb
             Nitrofenb
             PCNBb
             Perthaneb
             Propachlorb
             Strobaneb
             tra/?s-Nonachlorb
             trans-Permethrinb
             Trifluralinb
 15972-60-8
  2425-06-1
   133-06-2
  2675-77-6
 99516-95-7
  1897-45-6
  1861-32-1
   117-80-6
   115-32-2
  2593-15-9
 58718-66-4
 58718-67-5
 12616-35-2
 12616-36-3
  2234-13-1
 39450-05-0
  2385-85-5
  1836-75-5
    82-68-8
    72-56-0
 1918-16-17
  8001-50-1
 39765-80-5
 51877-74-8
  1582-09-8
                Single-column analysis
                Dual-column analysis
                                   8081 - 3
                          Revision 0
                      September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   A measured volume or weight of sample (approximately 1 L for 1 iquids,
2 g to  30  g for solids) is extracted  using  the  appropriate sample  extraction
technique.   Liquid samples are extracted at neutral pH with methylene chloride
using either  a  separatory  funnel  (Method 3510) or  a  continuous liquid-liquid
extractor (Method 3520). Solid  samples are extracted with hexane-acetone (1:1)
or methylene  chloride-acetone  (1:1)  using  either Soxhlet  extraction  (Method
3540), Automated Soxhlet (Method 3541), or Ultrasonic Extraction  (Method 3550).
A variety of cleanup steps may be applied to the extract, depending  on (1) the
nature  of  the coextracted matrix  interferences  and (2) the  target analytes.
After cleanup,  the extract  is  analyzed by injecting a 1-juL sample  into  a gas
chromatograph with  a narrow- or wide-bore  fused silica capillary  column and
electron capture  detector  (GC/ECD)  or an electrolytic  conductivity detector
(GC/ELCD).


3.0   INTERFERENCES

      3.1   Refer to Methods 3500 (Sec. 3, in particular),  3600, and 8000.

      3.2   Sources  of  interference  in this method can  be  grouped  into  three
broad categories:  contaminated solvents, reagents or sample processing hardware;
contaminated GC  carrier gas, parts,  column surfaces or detector surfaces; and the
presence of coeluting  compounds in  the sample matrix  to  which the  ECD will
respond. Interferences coextracted  from the  samples will  vary considerably from
waste to waste.   While general cleanup techniques are referenced  or provided as
part of this method,  unique samples may require additional cleanup approaches to
achieve desired degrees of discrimination and quantitation.

      3.3   Interferences  by   phthalate  esters   introduced   during   sample
preparation  can pose  a major  problem  in   pesticide  determinations.    These
materials  may be removed  prior to  analysis using  Gel  Permeation  Cleanup  -
pesticide  option  (Method  3640)  or as  Fraction  III of the  silica gel  cleanup
procedure  (Method 3630).  Common flexible plastics contain varying  amounts of
phthalate esters which are easily extracted or leached from such materials during
laboratory operations.  Cross-contamination  of clean glassware routinely occurs
when plastics  are handled during  extraction steps, especially when  solvent-wetted
surfaces are handled.  Interferences  from phthalate esters can best be minimized
by avoiding contact  with  any plastic  materials and checking  all solvents and
reagents for phthalate contamination.  Exhaustive cleanup of solvents, reagents
and  glassware  may  be  required  to  eliminate  background  phthalate  ester
contamination.

      3.4   Glassware must be scrupulously cleaned.  Clean all glassware as soon
as possible after use by  rinsing  with the last solvent  used.   This should be
followed by detergent washing with  hot  water,  and rinses  with  tap water and
organic-free reagent water.   Drain  the glassware and dry  in an oven at 130°C for
several hours or rinse with methanol  and drain.  Store dry glassware  in a clean
environment.

      3.5   The presence of  elemental sulfur will result  in  broad  peaks that
interfere with the detection of early-eluting organochlorine pesticides.  Sulfur


                                   8081  - 4                         Revision 0
                                                                 September 1994

-------
contamination should be expected with sediment  samples.  Method 3660  is suggested
for removal  of  sulfur.   Since the recovery of  Endrin  aldehyde (using  the TBA
procedure) is drastically  reduced,  this compound must  be  determined  prior to
sulfur cleanup.

      3.6   Waxes, lipids, and other high molecular weight co-extractables can
be removed by Gel-Permeation Cleanup (Method 3640).

      3.7    It  may be  difficult to  quantitate Aroclor  patterns  and  single
component  pesticides  together.   Some  pesticides  can  be removed  by  sulfuric
acid/permanganate cleanup (Method 3665)  and  silica  fractionation (Method 3630).
Guidance on the identification of PCBs is given in Sec. 7.

      3.8   The following target  analytes coelute using single column analysis:

            DB 608      Trifluralin/Dial late isomers
                        PCNP/Dichlone/Isodrin
                        DDD/Endosulfan II

            DB 1701     Captan/Chlorobenzilate
                        Captafol/Mirex
                        DDD/Endosulfan II
                        Methoxychlor/Endosulfan sulfate

            3.8.1 Other  halogenated pesticides  or  industrial chemicals  may
      interfere  with   the   analysis  of   pesticides.     Certain   co-eluting
      organophosphorus  pesticides   are  eliminated   by  the  Gel   Permeation
      Chromatography  cleanup -  pesticide  option  (Method  3640).    Co-eluting
      chlorophenols are  eliminated by Silica gel  (Method 3630),  Florisil (Method
      3620),  or Alumina (Method 3610) cleanup.

      3.9   The following compounds  coelute  using the dual  column analysis.  Two
temperature programs are provided for the same pair of columns as  option 1 and
option 2 for dual column analysis.  In general,  the DB-5 column resolves fewer
compounds that the DB-1701:

            3.9.1 DB-5/DB-1701, thin film,  slow ramp:  See Sec. 7  and Table 6.

                  DB-5  trans-Permethrin/Heptachlor epoxide
                        Endosulfan I/a-Chlordane
                        Perthane/Endrin
                        Endosulfan II/Chloropropylate/Chiorobenzi1 ate
                        4,4'-DDT/Endosulfan sulfate
                        Methoxychlor/Di cofol

            Perthane/Endrin  and  Chiorobenzilate/Endosulfan  II/Chloropropylate
      will  also  co-elute  on  DB-5  after  moderate  deterioration  in  column
      performance.
                                   8081 - 5                         Revision 0
                                                                September 1994

-------
                  DB-1701     Chlorothalonil/B-BHC
                              (5-BHC/DCPA/trans-Permethrin
                              or-Chlordane/trans-Nonachlor
                              Captan/Dieldrin
                              Chiorobenzi1 ate/Chioropropylate

            Chlorothalonil/B-BHC and or-Chlordane/trans-Nonachlor will co-elute
      on the DB-1701 column after moderate deterioration in column performance.

            Nitrofen,  Dichlone,  Carbophenothion,  Dichloran  and Kepone  were
      removed from  the composite mixture  because of extensive peak tailing on
      both  columns.    Simazine  and Atrazine  give poor  responses on  the ECD
      detector.  Triazine  compounds should  be  analyzed using Method 8141 (NPD
      option).

            3.9.2 DB-5/DB-1701, thick film,  fast ramp:   See Sec.  7 and Table 7.

                  DB-5        Diallate/a-BHC
                              Perthane/Endosulfan II
                              Chiorobenzi1 ate/Chioropropylate
                              Endrin/Nitrofen
                              4,4'-DDT/Endosulfan sulfate
                              Methoxychlor/Di colfol

                  DB-1701     or-Chl ordane/trans-Nonachl or (partially resolved)
                              4,4'-DDD/Endosulfan II (partially resolved)


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph:   an   analytical  system  complete   with  gas
chromatograph  suitable for  on-column  and  split-splitless   injection  and all
required accessories  including syringes, analytical columns,  gases, electron
capture detectors (ECD), and recorder/integrator or data system.

      The  columns   listed  in this  section  were  used   to  develop  the  method
performance data.  Their specification is not intended to prevent laboratories
from  using  columns that  are  developed after  promulgation  of the  method.
Laboratories may use other  capillary columns if they document  method performance
data (e.g.  chromatographic resolution,  analyte  breakdown,  and MDLs) equal  to or
better than those provided with the method.

            4.1.1 Single-column Analysis:

                  4.1.1.1     Narrow-bore columns:

                        4.1.1.1.1   Column  1 - 30 m x 0.25 or 0.32 mm internal
                  diameter (ID) fused  silica capillary  column chemically bonded
                  with SE-54 (DB 5 or equivalent), 1 /im film thickness.

                        4.1.1.1.2   Column  2 -  30  m x  0.25  mm ID fused silica
                  capillary  column chemically  bonded  with   35  percent phenyl
                                   8081 - 6                         Revision 0
                                                                September 1994

-------
      methylpolysiloxane  (DB  608,  SPB 608,  or equivalent),  25 /urn
      coating thickness, 1 /zm film thickness.

            4.1.1.1.3   Narrow bore columns should be installed in
      split/splitless (Grob-type) injectors.

      4.1.1.2     Wide-bore columns

            4.1.1.2.1   Column 1 - 30 m  x  0.53 mm ID fused silica
      capillary  column  chemically bonded  with 35  percent phenyl
      methylpolysiloxane (DB 608, SPB 608,  RTx-35, or equivalent),
      0.5 urn or  0.83 jum film  thickness.

            4.1.1.2.2   Column 2 - 30 m  x  0.53 mm ID fused silica
      capillary  column  chemically bonded  with 50  percent phenyl
      methylpolysiloxane  (DB  1701,  or  equivalent),  1.0  jum  film
      thickness.

            4.1.1.2.3   Column 3 - 30 m  x  0.53 mm ID fused silica
      capillary  column  chemically bonded with  SE-54 (DB 5, SPB 5,
      RTx5, or equivalent), 1.5 /xm film thickness.

            4.1.1.2.4   Wide-bore columns should be installed in 1/4
      inch injectors, with deactivated liners designed specifically
      for use with these columns.

4.1.2 Dual Column Analysis:

      4.1.2.1     Column pair 1:

            4.1.2.1.1   J&W Scientific press-fit Y-shaped glass 3-
      way union splitter (J&W Scientific, Catalog no. 705-0733) or
      Restek Y-shaped fused-silica connector  (Restek,  Catalog no.
      20405), or equivalent.

            4.1.2.1.2   30 m  x  0.53 m  ID  DB-5  (J&W Scientific),
      1.5 /Lim film thickness, or equivalent.

            4.1.2.1.3   30 m x 0.53 mm ID DB-1701 (J&W Scientific),
      1.0 jum film thickness, or equivalent.

      4.1.2.2     Column pair 2:

            4.1.2.2.1   Splitter 2 -  Supelco 8 in. glass injection
      tee,   deactivated  (Supelco,   Catalog   no.   2-3665M),   or
      equivalent.

            4.1.2.2.2   30 m  x  0.53 m  ID  DB-5  (J&W Scientific),
      0.83 /Ltm film thickness, or equivalent.

            4.1.2.2.3   30 m x 0.53 mm ID DB-1701 (J&W Scientific),
      1.0 jum film thickness, or equivalent.
                       8081 - 7                         Revision 0
                                                    September 1994

-------
            4.1.3 Column  rinsing  kit:   Bonded-phase  column  rinse  kit  (J&W
      Scientific, Catalog no. 430-3000 or equivalent).

      4.2   Glassware  (see  Methods  3510,  3520,  3540,  3541, 3550,  3630,  3640,
3660, and 3665 for specifications).

      4.3   Kuderna-Danish (K-D) apparatus. See extraction methods for specifics.


5.0   REAGENTS

      5.1   Reagent or  pesticide  grade  chemicals shall  be used  in all  tests.
Unless otherwise indicated,  it is  intended  that  all reagents  shall conform to
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where such  specifications  are  available.   Other grades  may  be  used,
provided it is first ascertained that the reagent  is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      NOTE: Store  the  standard   solutions  (stock,   composite,  calibration,
            internal, and surrogate) at 4°C in Teflon-sealed  containers in the
            dark.  When a lot of standards is prepared,  it is  recommended that
            aliquots of that lot be stored in individual  small  vials.  All  stock
            standard  solutions must  be replaced after  one  year  or  sooner if
            routine  QC (Sec.  8)   indicates  a problem.   All  other  standard
            solutions must be replaced  after six  months or sooner  if routine QC
            (Sec. 8) indicates a  problem.

      5.2   Solvents and reagents:  As appropriate for Method 3510, 3520,  3540,
3541, 3550,  3630,   3640,  3660,   or  3665:  n-hexane,  diethyl   ether,  methylene
chloride, acetone,  ethyl acetate, and isooctane (2,2,4-trimethylpentane).  All
solvents should  be  pesticide quality or  equivalent,  and each  lot  of solvent
should be determined to be phthalate free.  Solvents must  be exchanged to hexane
or isooctane prior to analysis.

            5.2.1 Organic-free reagent water:  All  references  to water in this
      method refer to organic-free reagent water as defined  in Chapter One.

      5.3   Stock standard  solutions  (1000 mg/L): Can  be prepared  from pure
standard materials or can be purchased as certified solutions.

            5.3.1 Prepare stock standard solutions by accurately weighing about
      0.0100 g of pure compound.   Dissolve the compound in isooctane or hexane
      and dilute to volume in a 10-mL volumetric flask.   If compound purity is
      96 percent or  greater, the weight  can  be used  without correction  to
      calculate the concentration of the stock standard solution.  Commercially
      prepared stock standard solutions  can be used  at  any concentration if they
      are certified by the manufacturer or by an independent  source.

            5.3.2 6-BHC,  Dieldrin,  and  some  other  standards  may  not  be
      adequately soluble  in  isooctane.   A small  amount  of acetone or toluene
      should be used to dissolve  these compounds during the  preparation of the
      stock standard solutions.
                                   8081 - 8                         Revision 0
                                                                September 1994

-------
      5.4   Composite stock  standard:  Can  be  prepared from  individual  stock
solutions.  For composite  stock  standards  containing  less than 25 components,
take exactly 1 ml of each individual stock solution at 1000 mg/L, add solvent,
and mix the solutions in  a 25-mL volumetric flask.  For example, for a composite
containing 20  individual standards, the resulting concentration of  each component
in the mixture,  after the volume  is  adjusted to 25 ml,  will be  1 mg/25 ml.  This
composite solution can be further diluted to obtain the desired concentrations.
For composite  stock standards containing more than 25 components,  use volumetric
flasks of the appropriate volume  (e.g.,  50 ml, 100 ml).

      5.5   Calibration  standards  should  be  prepared  at a  minimum of  five
concentrations by  dilution  of  the  composite stock standard with isooctane or
hexane.    The concentrations  should  correspond  to   the expected  range  of
concentrations found in real  samples and  should bracket the linear range of the
detector.

            5.5.1 Although all  single component  analytes  can  be  resolved  on a
      new  35  percent  phenyl  methyl  silicone   column   (e.g.,  DB-608),   two
      calibration mixtures should be prepared  for the  single component analytes
      of this method.

            5.5.2 This  procedure  is  established  to   (1) minimize  potential
      resolution and quantitation problems on confirmation columns or on older
      35  percent  phenyl  methyl  silicone  (e.g.  DB-608) columns  and  (2)  allow
      determination of Endrin and DDT breakdown for method QC (Sec.  8).

            5.5.3 Separate calibration standards  are  required for each  multi-
      component target analyte,  with the exception of Aroclors 1016 and 1260,
      which can be run as a mixture.

      5.6   Internal standard (optional):

            5.6.1 Pentachloronitrobenzene is suggested as an internal standard
      for the single column analysis, when it is not  considered to be a  target
      analyte.   l-Bromo-2-nitrobenzene  is  a  suggested  option.   Prepare  the
      standard to complement the concentrations found  in Sec.  5.5.

            5.6.2 Make a  solution of 1000 mg/L  of l-bromo-2-nitrobenzene  for
      dual-column  analysis.  Dilute it  to  500 ng/^L  for spiking,  then use a
      spiking volume of 10 /xL/mL  of extract.

      5.7   Surrogate  standards:   The  performance of the  method   should  be
monitored  using  surrogate compounds.   Surrogate standards  are  added  to  all
samples, method blanks, matrix spikes,  and calibration standards.

            5.7.1 For the single  column analysis,  use  decachlorobiphenyl  as the
      primary surrogate.  However, if recovery is  low,  or late-eluting compounds
      interfere with  decachlorobiphenyl,  then tetrachloro-m-xylene  should  be
      evaluated  as a  surrogate.    Proceed  with  corrective  action   when  both
      surrogates are out of limits  for a sample (Sec.  8.2).  Method  3500,  Sec.
      5, indicates the proper procedure  for preparing  these surrogates.
                                   8081 - 9                         Revision 0
                                                                September 1994

-------
            5.7.2 For the dual column analysis make a solution of 1000 mg/L of
      4-chloro-3-nitrobenzotrifluoride and dilute to 500 ng//iL.  Use a spiking
      volume of 100 pi  for  a 1  L aqueous sample.   Store the spiking solutions
      at 4°C in Teflon-sealed containers  in the dark.


6.0   SAMPLE COLLECTION, PRESERVATION AND HANDLING

      6.1   See Chapter 4, Organic Analytes, Sec.  4.

      6.2   Extracts must be stored under refrigeration in the dark and analyzed
within 40 days of extraction.


7.0   PROCEDURE

      7.1   Extraction:

            7.1.1 Refer to Chapter Two and Method  3500 for guidance in choosing
      the  appropriate  extraction procedure.    In  general,  water  samples are
      extracted at  a neutral pH with methylene chloride  as  a  solvent  using a
      separatory funnel  (Method  3510)  or a continuous liquid-liquid extractor
      (Method 3520).  Extract solid  samples with hexane-acetone  (1:1) using one
      of the Soxhlet extraction  (Method 3540 or 3541) or ultrasonic extraction
      (Method 3550) procedures.

            NOTE: Hexane/acetone  (1:1) may  be  more effective as an extraction
                  solvent  for  organochlorine  pesticides  and   PCBs  in  some
                  environmental   and  waste   matrices   than    is   methylene
                  chloride/acetone  (1:1).    Use  of hexane/acetone  generally
                  reduces the amount of co-extracted interferences and improves
                  signal/noise.

            7.1.2 Spiked samples  are used to verify the  applicability  of the
      chosen extraction  technique to each new  sample type.   Each  sample type
      must  be  spiked with  the compounds of interest  to  determine  the  percent
      recovery and the  limit of detection for  that sample  (Sec.  5).  See Method
      8000  for guidance on demonstration of initial method proficiency as well
      as guidance on matrix  spikes for routine  sample analysis.

      7.2   Cleanup/Fractionation:

            7.2.1 Cleanup procedures may  not be necessary for a relatively  clean
      sample matrix, but most extracts from environmental  and waste samples will
      require  additional  preparation before  analysis.    The  specific  cleanup
      procedure used will depend on the nature of  the sample  to  be analyzed and
      the data  quality  objectives for the measurements.   General guidance for
      sample extract cleanup  is provided in this section and in Method 3600.

                  7.2.1.1     If  a sample is of biological origin,  or contains
            high molecular  weight materials,  the  use  of GPC cleanup/pesticide
            option  (Method  3640) is  recommended.   Frequently,  one  of the
                                   8081  -  10                        Revision 0
                                                                September 1994

-------
            adsorption chromatographic cleanups may also be required following
            the GPC cleanup.

                  7.2.1.2     If only PCBs are to be measured in a sample, the
            sulfuric  acid/permanganate  cleanup (Method 3665)  is recommended.
            Additional cleanup/fractionation by Alumina Cleanup  (Method 3610),
            Silica-Gel Cleanup (Method 3630), or Florisil Cleanup  (Method 3620),
            may be necessary.

                  7.2.1.3     If both PCBs and  pesticides  are to  be measured in
            the sample, isolation of the PCB fraction by Silica Cleanup (Method
            3630) is recommended.

                  7.2.1.4     If only pesticides  are to be measured, cleanup by
            Method 3620 or Method 3630 is recommended.

                  7.2.1.5     Elemental   sulfur,  which may  appear  in  certain
            sediments  and  industrial  wastes,  interferes  with  the  electron
            capture gas chromatography of certain pesticides.  Sulfur should be
            removed by the technique described in Method 3660,  Sulfur Cleanup.

      7.3   GC Conditions:     This method allows the analyst to  choose between
a single column  or  a dual column configuration in  the  injector  port.   Either
wide- or narrow-bore  columns  may  be  used.   Identifications  based on retention
times from  a  single column  must  be confirmed on  a second column or  with an
alternative qualitative technique.

            7.3.1 Single Column Analysis:

                  7.3.1.1     This capillary GC/ECD method  allows  the  analyst
            the option of using  0.25-0.32 mm ID capillary columns (narrow-bore)
            or 0.53 mm ID capillary columns (wide-bore).   Performance data are
            provided   for  both  options.     Figures   1-6  provide   example
            chromatograms.

                  7.3.1.2     The use of  narrow-bore columns  is recommended when
            the analyst  requires  greater chromatographic  resolution.   Use of
            narrow-bore columns is suitable for relatively clean  samples or for
            extracts that have been  prepared with  one or  more  of the clean-up
            options referenced in the method.   Wide-bore columns (0.53  mm) are
            suitable for more complex environmental and waste matrices.

                  7.3.1.3     For the single column method  of  analysis,  using
            wide-bore capillary columns, Table 1  lists  average  retention times
            and method detection limits  (MDLs)  for  the target analytes in water
            and soil  matrices.  For  the  single  column method of analysis, using
            narrow-bore capillary columns, Table 2 lists average retention times
            and method detection limits  (MDLs)  for  the target analytes in water
            and soil  matrices. The MDLs for the components of a specific sample
            may differ  from  those listed in Tables 1  and 2 because they are
            dependent upon the  nature  of interferences in  the  sample  matrix.
            Table 3  lists  the Estimated Quantitation  Limits (EQLs)  for other
            matrices.  Table 4 lists  the  GC operating conditions  for the single
            column method of analysis.

                                  8081  -  11                         Revision 0
                                                                September 1994

-------
      7.3.2 Dual Column Analysis:

            7.3.2.1     The dual-column/dual-detector  approach  involves
      the use of two 30 m x 0.53 mm ID fused-silica open-tubular columns
      of different polarities,  thus different  selectivities  towards the
      target compounds. The columns are connected to an  injection tee and
      ECD detectors.  Retention times  for the organochlorine analytes on
      dual columns are in Table  5.  The  GC  operating conditions for the
      compounds in Table  5  are in Table 6.  Multicomponent  mixtures of
      Toxaphene and Strobane were  analyzed  separately  (Figures  7  and 8)
      using the GC operating conditions found in Table 7.   Seven Aroclor
      mixtures and six  Halowax mixtures were analyzed under the conditions
      outlined in Table 7 (Figures 9 through 21).  Figure 22 is a sample
      chromatogram  for a  mixture  of  organochlorine  pesticides.    The
      retention times  of the  individual components detected  in  these
      mixtures are given in Tables 8 and 9.

                  7.3.2.1.1   Operating  conditions   for  a more  heavily
            loaded DB-5/DB-1701 pair are given  in Table 7.   This column
            pair   was   used   for  the   detection   of  multi component
            organochlorine compounds.

                  7.3.2.1.2   Operating  conditions   for a  DB-5/DB-1701
            column pair with thinner films,  a different  type of splitter,
            and a  slower temperature  programming  rate are  provided in
            Table  6.    These  conditions gave  better  peak  shapes  for
            compounds  such  as  Nitrofen and  Dicofol.  Table  5  lists the
            retention  times  for the compounds  detected on  this  column
            pair.

7.4   Calibration:

      7.4.1 Prepare calibration standards using  the  procedures in Sec. 5.
Refer to Method 8000 (Sec. 7) for proper calibration techniques for both
initial  calibration  and  calibration  verification.    The procedure  for
either  internal or external calibration may  be used,  however,  in most
cases external  standard  calibration  is used with Method 8081.   This is
because  of the sensitivity  of  the  electron capture  detector and  the
probability of  the internal standard  being affected  by  interferences.
Because  several  of the  pesticides may  co-elute  on any  single column,
analysts should use two calibration mixtures (see Sec. 3.8).  The specific
mixture should be selected to minimize the problem of peak overlap.

      NOTE: Because of the sensitivity of the electron capture detector,
            the injection port and column should always be cleaned prior
            to performing the initial  calibration.

            7.4.1.1     Method  8081  has  many  multi-component  target
      analytes.    For  this reason,   the   target   analytes  chosen  for
      calibration  should  be limited to  those  specified in  the project
      plan.   For  instance,  some  sites  may  require  analysis for  the
      organochlorine pesticides only or the PCBs only.   Toxaphene and/or
      technical Chlordane  may  not  be  specified  at certain  sites.   In
      addition, where PCBs are  specified in the project  plan,  a mixture of

                             8081  - 12                         Revision 0
                                                          September  1994

-------
      Aroclors 1016 and 1260 will suffice for the initial calibration of
      all  Aroclors,  since  they include  all  congeners  present  in  the
      different regulated Aroclors.  A mid-point calibration standard of
      all Aroclors (for Aroclor pattern recognition)  must  be included with
      the  initial calibration so that  the  analyst  is familiar with each
      Aroclor pattern and retention times on each column.

            7.4.1.2     For calibration verification  (each  12 hr shift)
      all  target  analytes required  in  the  project  plan must  be injected
      with  the  following exception  for  the Aroclors.   For  sites  that
      require PCB analysis,  include only the Aroclors that are expected to
      be found at the site.  If PCBs are required, but it  is unknown which
      Aroclors may be  present,  the  mid-concentration Aroclors 1016/1260
      mixture only, may  be  injected.   However,  if  specific  Aroclors are
      found at the site during the  initial  screening,  it  is required that
      the  samples containing Aroclors be reinjected with the proper mid-
      concentration Aroclor standards.

      7.4.2 Because  of  the  low  concentration   of   pesticide  standards
injected on a GC/ECD,  column adsorption may be a problem when the GC has
not been  used for a day  or more.    Therefore,  the GC column  should be
primed or  deactivated  by injecting a PCB  or  pesticide standard mixture
approximately  20 times  more  concentrated  than  the  mid-concentration
standard.   Inject this standard mixture prior to  beginning  the initial
calibration or calibration verification.

      CAUTION:    Several analytes, including Aldrin, may be observed in
                  the  injection just  following  this   system  priming.
                  Always run  an acceptable  blank  prior to  running  any
                  standards or samples.

      7.4.3 Retention time windows:

            7.4.3.1     Before establishing the  retention time windows,
      make sure the gas chromatographic system is within optimum operating
      conditions.  The  width of  the retention time window should be based
      upon actual  retention times of standards measured over the course of
      72 hours.   See Method 8000 for details.

            7.4.3.2     Retention time windows shall be defined as plus or
      minus three times the standard deviation of the absolute retention
      times for each standard.   However,  the experience of  the analyst
      should weigh heavily  in  the  interpretation of the chromatograms.
      For multicomponent standards (i.e.,  PCBs), the analyst  should  use
      the  retention  time window but should  primarily rely  on  pattern
      recognition.  Sec.  7.5.4 provides guidance on the establishment of
      absolute retention time  windows.

            7.4.3.3     Certain  analytes, particularly Kepone, are subject
      to changes  in retention times.   Dry Kepone standards  prepared in
      hexane or  isooctane can  produce  gaussian  peaks.   However,  Kepone
      extracted  from samples or standards exposed to  water or methanol  may
      produce peaks with broad tails that  elute  later than  the standard
      (0-1 minute).  This shift  is presumably the result of the formation

                            8081 - 13                         Revision 0
                                                          September 1994

-------
      of  a  hemi-acetal  from the  ketone  functionality.   Method  8270 is
      recommended for Kepone.

7.5   Gas chromatographic analysis:

      7.5.1 Set up the GC system using the conditions described in Tables
4, 6, or 7.  An initial oven temperature at or below 140-150°C is required
to resolve  the  four BHC  isomers.   A final  temperature  of 240-270°C is
required  to  elute  decachlorobiphenyl.     Use   of  injector  pressure
programming will  improve the chromatography of late eluting peaks.

      7.5.2 Verify calibration  each 12 hour shift  by injecting calibration
verification standards prior to conducting any analyses.   See Sec. 7.4.1.2
for special guidance on calibration verification  of PCBs.  A calibration
standard must also be  injected  at  intervals  of not less  than once every
twenty  samples  (after  every 10  samples  is  recommended  to  minimize  the
number of samples requiring  re-injection when QC  limits are exceeded)  and
at the  end  of the analysis  sequence.   The calibration  factor  for each
analyte to  be quantitated must  not exceed  a  ±15  percent  difference when
compared  to the  initial calibration  curve.    When  this  criterion  is
exceeded, inspect the gas chromatographic  system to determine the cause
and perform whatever maintenance is necessary  before verifying calibration
and  proceeding with  sample  analysis.   If routine  maintenance  does  not
return the  instrument performance to meet the QC  requirements (Sec. 8.2)
based on the last initial calibration, then a  new  initial calibration must
be performed.

            7.5.2.1     Analysts should use high and low concentrations of
      mixtures of single-component analytes and multi-component analytes
      for calibration verification.

      7.5.3 Sample injection may continue for as  long as  the calibration
verification standards and  standards  interspersed  with the samples meet
instrument QC  requirements.   It is  recommended that  standards be analyzed
after every 10 (required  after  every 20 samples),  and at the end of a set.
The  sequence  ends when  the set  of  samples  has been injected  or when
qualitative and/or quantitative QC criteria are exceeded.

            7.5.3.1     Each sample  analysis  must  be bracketed  with an
      acceptable initial calibration, calibration  verification standard(s)
      (each 12 hr shift), or calibration standards interspersed within the
      samples.   All  samples that were injected  after the  standard that
      last met the QC criteria must be reinjected.

            7.5.3.2     Although analysis of  a  single mid-concentration
      standard (standard mixture or multi-component analyte) will satisfy
      the  minimum  requirements,  analysts are   urged  to  use  different
      calibration    verification    standards    during    organochlorine
      pesticide/PCB analyses.   Also,  multi-level standards (mixtures or
      multi-component  analytes)  are highly  recommended  to  ensure that
      detector  response  remains  stable  for   all  analytes  over  the
      calibration range.
                             8081  -  14                         Revision 0
                                                          September 1994

-------
      7.5.4 Establish  absolute  retention  time  windows for each analyte.
Use the absolute retention time for each analyte from standards analyzed
during  that  12 hour  shift as  the  midpoint of  the  window.   The daily
retention  time window  equals  the midpoint  +  three  times  the standard
deviations.

            7.5.4.1     Tentative identification of an analyte occurs when
      a peak from a sample extract falls within the daily retention time
      window.

            7.5.4.2     Validation   of   gas   chromatographic   system
      qualitative  performance:  Use  the  calibration  standards analyzed
      during the sequence to evaluate retention time stability.  If any of
      the standards fall outside their daily retention time windows, the
      system is out  of control. Determine the  cause  of  the problem and
      correct  it.

      7.5.5 Record the volume  injected to  the  nearest  0.05  juL and the
resulting  peak size  in  area  units.    Using  either  the  internal  or the
external calibration procedure  (Method 8000), determine the identity and
the quantity  of each  component peak  in  the sample  chromatogram which
corresponds to the compounds used for calibration purposes.

            7.5.5.1     If the responses exceed the calibration range of
      the  system,  dilute  the   extract   and   reanalyze.   Peak  height
      measurements  are  recommended   over  peak area  integration  when
      overlapping peaks cause errors in area integration.

            7.5.5.2     If partially overlapping or  coeluting peaks are
      found, change  columns  or try  GC/MS  quantitation,  see  Sec.  8 and
      Method 8270.

            7.5.5.3     If the  peak response is  less  than  2.5 times the
      baseline noise  level, the validity of the  quantitative  result may be
      questionable.  The  analyst  should  consult with the  source  of the
      sample to determine whether further concentration of the sample is
      warranted.

      7.5.6 Identification of mixtures  (i.e.  PCBs and Toxaphene) is based
on  the  characteristic  "fingerprint"  retention  time and  shape  of  the
indicator  peak(s);  and  quantitation  is  based  on  the  area  under  the
characteristic  peaks  as compared  to the  area  under the  corresponding
calibration peak(s) of the same retention time  and  shape generated using
either internal or external calibration procedures.

      7.5.7 Quantitation of  the target compounds  is based  on:   1)   a
reproducible response of the ECD or ELCD within the calibration range; and
2)  a  direct  proportionality between  the  magnitude  of  response  of the
detector to peaks  in  the sample extract and the calibration standards.
Proper quantitation requires  the appropriate  selection of a baseline from
which  the area  or  height of the  characteristic peak(s) can be determined.
                             8081  -  15                         Revision 0
                                                          September 1994

-------
      7.5.8 If compound identification or quantitation is precluded due to
interference  (e.g.,  broad,  rounded  peaks  or ill-defined  baselines  are
present) cleanup of the extract  or  replacement of the capillary column or
detector  is  warranted.    Rerun the sample  on another  instrument  to
determine if  the problem  results from  analytical hardware  or the sample
matrix.  Refer to Method 3600  for the procedures to be followed in sample
cleanup.

7.6   Quantitation of Multiple Component Analytes:

      7.6.1 Multi-component analytes present  problems   in  measurement.
Suggestions are offered in the following sections for handing Toxaphene,
Chlordane, PCBs, DDT, and BHC.

      7.6.2 Toxaphene:  Toxaphene is manufactured by the chlorination of
camphenes, whereas Strobane results from the  chlorination of a mixture of
camphenes and pinenes.  Quantitative calculation of Toxaphene or Strobane
is difficult,  but reasonable  accuracy  can  be  obtained.   To  calculate
Toxaphene on  GC/ECD:   (a)    adjust  the sample  size  so that  the  major
Toxaphene peaks are  10-70% of full-scale deflection (FSD);  (b)   inject a
Toxaphene standard that is  estimated to be within  +10  ng of the sample;
(c)  quantitate  using  the five major  peaks  or  the total  area of  the
Toxaphene pattern.

            7.6.2.1     To measure total area,  construct the baseline of
      standard  Toxaphene  between   its  extremities;  and  construct  the
      baseline under the sample, using the distances of the peak troughs
      to  baseline  on the standard  as  a guide.   This procedure  is  made
      difficult by the  fact that the relative  heights  and  widths of the
      peaks in the sample will  probably  not be identical to the standard.

            7.6.2.2     A   series   of   Toxaphene  residues  have   been
      calculated using the total  peak area for comparison to the standard
      and also using the area  of the last four peaks only,  in both sample
      and standard.  The agreement between the results obtained  by the two
      methods justifies the  use  of the  latter method  for calculating
      Toxaphene  in  a  sample  where the  early  eluting  portion of  the
      Toxaphene  chromatogram  shows  interferences  from  other substances
      such as DDT.

      7.6.3 Chlordane  is  a  technical  mixture  of  at   least   11  major
components and 30 or more minor components.   Trans- and c/s-Chlordane (a
and 7, respectively), are the two major components of technical Chlordane.
However,  the  exact  percentage of each  in the  technical  material  is not
completely defined,  and is not consistent from batch to batch.

            7.6.3.1     The GC pattern of a Chlordane residue may differ
      considerably from that of the technical standard.   Depending on the
      sample substrate and its history,  residues  of Chlordane can consist
      of  almost any  combination  of:   constituents  from  the  technical
      Chlordane,  plant  and/or  animal  metabolites,  and  products  of
      degradation  caused  by exposure  to  environmental   factors  such as
      water and sunlight.


                            8081  - 16                         Revision 0
                                                          September 1994

-------
            7.6.3.2     Whenever possible, when a Chlordane residue does
      not resemble technical  Chlordane,  the analyst should quantitate the
      peaks of ff-Chlordane, 7-Chlordane, and Heptachlor separately against
      the  appropriate reference  materials,  and  report the  individual
      residues.

            7.6.3.3     When the GC pattern of the residue resembles that
      of  technical  Chlordane,   the  analyst  may  quantitate  Chlordane
      residues by comparing the total area of the Chlordane chromatogram
      using the  five  major  peaks or the total area.   If the Heptachlor
      epoxide peak is relatively  small,  include  it  as part  of the total
      Chlordane area for  calculation of the residue.   If Heptachlor and/or
      Heptachlor  epoxide are much out  of  proportion,  calculate  these
      separately and  subtract their areas from the  total  area to give a
      corrected  Chlordane  area.    (Note  that  octachloro  epoxide,  a
      metabolite  of  Chlordane,   can easily be  mistaken for  Heptachlor
      epoxide on a nonpolar GC column.)

            7.6.3.4     To  measure  the  total   area   of  the  Chlordane
      chromatogram, inject an amount of technical  Chlordane standard which
      will  produce   a   chromatogram   in  which   the   major  peaks  are
      approximately the same size as those in  the sample chromatograms.

      7.6.4 Polychlorinated biphenyls  (PCBs):  Quantitation of  residues of
PCBs involves  problems similar to those encountered in  the quantitation of
Toxaphene, Strobane,  and  Chlordane.  In each case, the  material is made up
of numerous compounds which generate multi-peak chromatograms.  Also, in
each case,  the  chromatogram of  the residue  may  not  .match  that  of  the
standard.

             7.6.4.1    Mixtures of PCBs of various chlorine  contents were
      sold for many years in  the  U.S.  by the Monsanto Co. under the trade
      name Aroclor (1200  series and  1016). Although these Aroclors are no
      longer  marketed,   the  PCBs   remain  in   the  environment  and  are
      sometimes found as  residues in foods,  especially fish.  The Aroclors
      most commonly found in the environment  are  1242, 1254, and 1260.

            7.6.4.2     PCB  residues   are   generally   quantitated   by
      comparison to the most similar Aroclor  standard.  A choice must be
      made as  to which Aroclor is most  similar to that of the residue  and
      whether that standard  is  truly  representative of the PCBs  in  the
      sample.

            7.6.4.3     PCB  Quantitation option  #1-  Quantitate  the  PCB
      residues by  comparing  the  total  area of the  chlorinated  biphenyl
      peaks to  the total  area   of  peaks from  the  appropriate  Aroclor
      reference material. Measure  the  total area or height response from
      the common baseline under  all  the peaks.  Use only those peaks from
      the sample that can be attributed to chlorobiphenyls.   These peaks
      must also  be  present in the chromatogram of the reference materials.
      Option #1  should not be used  if there are interference  peaks within
      the Aroclor pattern,  especially if they  overlap PCB congeners.
                            8081 - 17                         Revision 0
                                                          September 1994

-------
                  7.6.4.4     PCB Quantitation  option  #2- Quantitate  the  PCB
            residues by comparing the responses of 3 to  5 major peaks in each
            appropriate  Aroclor  standard  with the  peaks  obtained  from  the
            chlorinated biphenyls in  the  sample  extract.  The amount of Aroclor
            is calculated using  an individual  response factor  for  each of the
            major peaks.  The results of the 3 to 5 determinations are averaged.
            Major peaks are defined as those peaks  in the Aroclor standards that
            are at least 25% of the height of the  largest Aroclor peak.  Late-
            eluting  Aroclor  peaks   are   generally  the  most  stable  in  the
            environment.

                  7.6.4.5     When samples  appear  to  contain  weathered  PCBs,
            treated PCBs, or mixtures of  Aroclors, the use of Aroclor standards
            is not appropriate.  Several  diagnostic peaks useful for identifying
            non-Aroclor PCBs are  given  in Table 10.   Analysts  should examine
            chromatograms containing  these peaks carefully, as these samples may
            contain PCBs.   PCB concentrations may be estimated  from specific
            congeners by adding the concentration  of  the  congener peaks listed
            in Table 11.  The congeners are analyzed as single components.  This
            approach will provide reasonable accuracy  for Aroclors  1016,  1232,
            1242 and 1248 but will underestimate the concentrations of Aroclors
            1254,  1260  and  1221.   It  is  highly recommended  that  heavily
            weathered,  treated,  or mixed Aroclors be  analyzed  using  GC/MS if
            concentration permits.

            7.6.5 Hexachlorocyclohexane   (BHC,  from the  former  name,  benzene
      hexachloride):   Technical  grade  BHC is a cream-colored  amorphous  solid
      with a very  characteristic musty  odor; it consists of a  mixture of six
      chemically distinct isomers and one  or more  heptachlorocyclohexanes and
      octachlorocyclohexanes.    Commercial  BHC  preparations may  show a  wide
      variance in  the  percentage of  individual isomers present.    Quantitate
      each  isomer (a,  |8,  7,  and  5)  separately  against  a  standard of  the
      respective pure isomer.

            7.6.6 DDT:   Technical DDT consists primarily  of a mixture of 4,4'-
      DDT  (approximately 75%)   and  2,4'-DDT  (approximately  25%).    As  DDT
      weathers, 4,4'-DDE, 2,4'-DDE,  4,4'-DDD,  and  2,4'-DDD  are formed.  Since
      the 4,4'-isomers  of  DDT, DDE,  and ODD predominate in the environment,
      these  are   the  isomers  normally   regulated  by  US  EPA   and  should  be
      quantitated against standards of the respective  pure isomer.

      7.7   Suggested chromatography maintenance: Corrective measures may require
any one or more of the following remedial actions.

            7.7.1 Splitter connections:   For  dual  columns which  are  connected
      using a  press-fit Y-shaped  glass  splitter  or  a Y-shaped  fused-silica
      connector (J&W Scientific  or Restek),  clean  and  deactivate the splitter
      port insert or replace with a cleaned and deactivated splitter.  Break off
      the first few inches (up to one foot)  of  the injection port  side of the
      column.    Remove  the  columns  and  solvent  backflush  according to  the
      manufacturer's instructions. If these procedures fail to  eliminate the
      degradation problem,  it  may be  necessary to deactivate the metal injector
      body and/or replace the columns.


                                  8081 - 18                         Revision 0
                                                                September 1994

-------
                  7.7.1.1     GC  injector ports  can  be of  critical  concern,
            especially  in  the  analysis  of DDT and Endrin.   Injectors that are
            contaminated,   chemically   active,   or   too  hot  can  cause  the
            degradation ("breakdown") of the analytes.  Endrin and  DDT breakdown
            to Endrin aldehyde, Endrin ketone, ODD, or DDE.  When such breakdown
            is observed, clean  and  deactivate the injector port,  break off at
            least  0.5 M of the column  and  remount  it.   Check  the injector
            temperature  and lower  it to  205°C,  if  required.   Endrin  and DDT
            breakdown is less of a problem when  ambient  on-column  injectors are
            used.

            7.7.2 Metal  injector body:    Turn  off  the oven  and  remove the
      analytical columns when the oven has cooled.  Remove the glass injection
      port insert (instruments with on-column injection).  Lower  the injection
      port temperature  to  room  temperature.   Inspect  the injection  port and
      remove any noticeable foreign material.

                  7.7.2.1     Place a beaker  beneath  the injector port inside
            the oven.  Using a wash  bottle, serially rinse  the entire inside of
            the injector port with  acetone and then toluene;  catch the rinsate
            in the beaker.

                  7.7.2.2     Prepare a  solution of a deactivating  agent (Sylon-
            CT or equivalent)  following  manufacturer's  directions.   After all
            metal surfaces  inside the injector body have been thoroughly coated
            with  the  deactivation  solution,  rinse  the  injector  body  with
            toluene, methanol,  acetone,  then  hexane.   Reassemble  the injector
            and replace the columns.

            7.7.3 Column rinsing:   The  column  should  be  rinsed  with  several
      column  volumes of  an  appropriate  solvent.    Both  polar  and  nonpolar
      solvents are recommended.  Depending on  the  nature of the sample residues
      expected,  the  first  rinse might  be  water,  followed  by  methanol  and
      acetone; methylene chloride is a good final rinse and  in some cases may be
      the  only solvent  required.    The  column  should then  be  filled  with
      methylene  chloride  and  allowed  to stand  flooded  overnight to  allow
      materials within  the  stationary phase  to  migrate  into  the solvent.   The
      column is then flushed with fresh methylene chloride, drained, and dried
      at room temperature with a stream of ultrapure nitrogen.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for  specific quality control  (QC)  procedures
including matrix spikes, duplicates and blanks.  Quality  control  to validate
sample  extraction  is  covered  in  Method 3500  and  in  the  extraction  method
utilized.  If an extract cleanup was performed, follow the QC  in Method 3600 and
in the specific cleanup method.

      8.2   Quality control requirements for the GC system, including calibration
and corrective  actions,  are found  in  Method 8000.   The following  steps are
recommended as additional method QC.
                                  8081  -  19                         Revision 0
                                                                September 1994

-------
            8.2.1 The  QC  Reference  Sample  concentrate  (Method 3500)  should
      contain the organochlorine pesticides at 10 mg/L for water  samples.   If
      this  method  is  to  be  used  for analysis  of  Aroclors,  Chlordane,  or
      Toxaphene  only,  the  QC  Reference  Sample  should   contain  the  most
      representative multi-component mixture at a concentration of 50 mg/L in
      acetone.  The frequency of analysis of the QC reference sample analysis is
      equivalent to  a minimum of 1 per 20 samples  or  1  per batch if less than 20
      samples.  If the recovery of any compound found in the QC reference sample
      is less  than  80 percent  or  greater than  120 percent  of the  certified
      value, the laboratory  performance is  judged to be out of control, and the
      problem must  be corrected.  A new set of  calibration  standards should be
      prepared and  analyzed.

            8.2.2 Calculate surrogate standard recovery on all samples, blanks,
      and  spikes.     Determine if   the  recovery   is  within  limits  (limits
      established by performing QC procedures outlined in Method  8000).

            If recovery is not within limits, the following are required:

                  8.2.2.1      Confirm that there  are no errors in calculations,
            surrogate solutions and  internal  standards.  Also, check instrument
            performance.

                  8.2.2.2      Examine chromatograms for interfering  peaks  and
            for integrated areas.

                  8.2.2.3      Recalculate the data and/or reanalyze the extract
            if any  of the  above checks reveal a  problem.

                  8.2.2.4      Reextract and reanalyze  the sample if none of the
            above are a problem or flag the data  as  "estimated concentration."

            8.2.3 Include  a calibration standard  after each group of 20 samples
      (it is recommended that  a calibration  standard be included after every 10
      samples  to minimize the  number of  repeat  injections)  in  the  analysis
      sequence as a calibration check.  The response  factors for the calibration
      should  be within 15  percent  of the  initial   calibration.   When  this
      continuing calibration is out  of  this  acceptance window,  the laboratory
      should stop analyses and take corrective  action.

            8.2.4 Whenever  quantitation  is accomplished  using  an  internal
      standard,  internal  standards  must  be evaluated  for  acceptance.   The
      measured  area  of the  internal  standard must be no more  than 50 percent
      different from the average area calculated  during  calibration.  When the
      internal  standard peak area is  outside the  limit, all  samples that  fall
      outside the QC criteria must be reanalyzed.

      8.3   DDT and  Endrin are easily degraded in  the injection port.  Breakdown
occurs when the injection  port liner is contaminated high  boiling residue  from
sample  injection  or  when  the  injector  contains  metal  fittings.   Check  for
degradation problems  by injecting a standard containing only  4,4'-DDT and Endrin.
Presence of 4,4'-DDE, 4,4'-ODD,  Endrin ketone or Endrin indicates breakdown.  If
degradation of either DDT  or Endrin exceeds 15%,  take  corrective action before
proceeding with calibration.

                                  8081  -  20                         Revision 0
                                                                September  1994

-------
            8.3.1 Calculate percent breakdown as follows:

            % breakdown    Total DDT degradation peak area  (DDE + ODD)
            for 4, 4' -DDT = -  x  100
                                    peak areas  (DDT + DDE + ODD)

                              Total endrin degradation peak area
            % breakdown       (Endrin aldehyde  + Endrin ketone)
            for Endrin   =     -       x 100
                              peak areas (Endrin + aldehyde +  ketone)

            8.3.2 The  breakdown  of DDT  and  Endrin should  be  measured before
      samples are analyzed and at the beginning  of each 12 hour  shift.  Injector
      maintenance  and  recalibration should  be  completed if the  breakdown is
      greater than  15% for either compound (Sec. 8.2.3).

      8.4   GC/MS  confirmation  may be  used  for  single  column analysis.   In
addition,  any compounds  confirmed  by two columns  should  also  be  confirmed by
GC/MS if the concentration is sufficient for detection by GC/MS.

            8.4.1 Full -scan GC/MS will  normally  require a minimum concentration
      near  10 ng/jLtL in the final extract  for  each single-component compound.
      Ion  trap  or  selected   ion monitoring  will  normally  require  a  minimum
      concentration  near 1 ng/jul_.

            8.4.2 The  GC/MS  must  be  calibrated  for  the   specific  target
      pesticides when it is used for quantitative analysis.

            8.4.3 GC/MS may  not  be used  for single  column  confirmation  when
      concentrations are below 1
            8.4.4 GC/MS  confirmation  should be accomplished  by analyzing the
      same extract used  for GC/ECD analysis and the associated  blank.

            8.4.5 Use of the base/neutral -acid extract and  associated  blank may
      be used if the surrogates and internal standards do not interfere and it
      is demonstrated that the analyte is  stable during acid/base partitioning.
      However,  if the  compounds are  not  detected  in  the  base/neutral -acid
      extract even though the concentrations are high enough, a GC/MS analysis
      of the pesticide extract should be performed.

            8.4.6 A QC reference  sample of the compound must also be  analyzed by
      GC/MS.  The concentration  of  the  QC reference  standard must demonstrate
      the ability to confirm the pesticides/Aroclors identified by GC/ECD.

      8.5   Whenever silica gel  (Method  3630)  or Florisil (Method 3620) cleanup
is  used,   the  analyst   must  demonstrate  that  the   fractionation  scheme  is
reproducible.   Batch  to  batch variation in the composition  of  the  silica gel
material or  overloading  the  column may  cause a  change in  the distribution
patterns of the  organochlorine pesticides and PCBs. When  compounds are found in
two fractions, add the concentrations  in the fractions, and corrections for any
additional  dilution.
                                   8081  -  21                         Revision 0
                                                                September 1994

-------
9.0   METHOD PERFORMANCE

      9.1   The MDL is  defined  in Chapter One.  The MDL concentrations listed in
Tables 1 and  2  were obtained using organic-free reagent water  and  sandy loam
soil.

      9.2   The chromatographic separations in this method have been tested in
a single laboratory by using clean  hexane  and  liquid  and  solid waste extracts
that were  spiked with the  test compounds  at  three concentrations.   Single-
operator precision,  overall  precision, and  method  accuracy were found  to be
related to the concentration of the compound and the type of matrix.

      9.3   This method has  been applied in a variety of commercial laboratories
for environmental  and  waste matrices.   Performance data were  obtained  for a
limited number of target analytes spiked into sewage sludge and dichloroethene
still  bottoms at high concentration levels.  These data are provided in Tables
12 and 13.

      9.4   The accuracy and precision obtainable with this  method depend on the
sample matrix,  sample  preparation  technique,  optional  cleanup techniques,  and
calibration procedures used.

      9.5   Single  laboratory  accuracy data were obtained  for organochlorine
pesticides  in  a clay  soil.   The  spiking  concentration was  500  M9/kg.   The
spiking solution was mixed into the soil and then immediately transferred to the
extraction device and  immersed  in the extraction solvent.  The  spiked sample was
then extracted by Method  3541 (Automated  Soxhlet).   The data represent a single
determination.   Analysis  was by capillary  column  gas  chromatography/electron
capture detector following Method 8081 for the organochlorine pesticides.  These
data are listed in Table 14 and were taken from Reference 14.

      9.6   Single laboratory recovery data were obtained for PCBs in clay and
soil.     Oak  Ridge  National  Laboratory  spiked  Aroclors   1254 and  1260  at
concentrations  of  5 and  50 ppm into  portions of  clay  and  soil  samples  and
extracted  these  spiked samples using  the  procedure outlined  in  Method  3541.
Multiple extractions using two different extractors were performed.  The extracts
were analyzed by Method 8081.  The  data  are listed  in  Table 15 and were taken
from Reference 15.

      9.7   Multi-laboratory accuracy and precision  data were  obtained for PCBs
in soil.   Eight laboratories spiked Aroclors 1254 and 1260 into three portions
of 10 g of Fuller's Earth on three non-consecutive days, followed by immediate
extraction using Method 3541.  Six  of the laboratories spiked  each Aroclor at 5
and 50 mg/kg and two laboratories  spiked  each Aroclor at 50 and  500 mg/kg.  All
extracts were analyzed by Oak  Ridge National  Laboratory,  Oak  Ridge, TN,  using
Method 8081.  These  data are listed in Table  16 and were taken from Reference 13.
                                   8081  -  22                         Revision 0
                                                                September 1994

-------
10.0  REFERENCES

1.    Lopez-Avila, V.;  Baldin,  E.;  Benedicto,  J; Milanes, J.;  Beckert.  W. F.
      Application of Open-Tubular Columns to SW-846 GC Methods";  final report to
      the U.S.  Environmental Protection Agency  on Contract  68-03-3511;  Mid-
      Pacific Environmental  Laboratory, Mountain View, CA, 1990.

2.    Development and Application of Test Procedures for Specific Organic Toxic
      Substances in Wastewaters.  Category  10  -  Pesticides  and  PCB Report for
      the U.S. Environmental Protection Agency on Contract 68-03-2606.

3.    Goerlitz, D.F.; Law,  L.M.  "Removal of Elemental  Sulfur Interferences from
      Sediment Extracts for Pesticide Analysis"; Bull.  Environ. Contam. Toxicol.
      1971,  6, 9.

4.    Ahnoff, M.; Josefsson, B.  "Cleanup Procedures for PCB Analysis on River
      Water Extracts"; Bull. Environ. Contam. Toxicol. 1975,  13, 159.

5.    Jensen, S.; Renberg, L.; Reutergardth, L.  "Residue Analysis of Sediment
      and Sewage Sludge  for  Organochlorines  in  the  Presence of  Elemental
      Sulfur"; Anal. Chem. 1977, 49, 316-318.

6.    Wise,  R.H.; Bishop, D.F.; Williams,  R.T.; Austern, B.M.   "Gel Permeation
      Chromatography  in the GC/MS  Analysis  of  Organics  in  Sludges";  U.S.
      Environmental  Research Laboratory.  Cincinnati, OH  45268.

7.    Pionke, H.B.;  Chesters, G.; Armstrong, D.E.   "Extraction  of Chlorinated
      Hydrocarbon Insecticides from Soil";  Agron. J. 1968, 60, 289.

8.    Burke, J.A.; Mills, P.A.; Bostwick, D.C.  "Experiments with Evaporation of
      Solutions of Chlorinated Pesticides"; J. Assoc. Off.  Anal.  Chem. 1966, 49,
      999.

9.    Glazer, J.A.,  et al.   "Trace Analyses  for Wastewaters"; Environ. Sci. and
      Technol. 1981, 15, 1426.

10.   Marsden, P.O.,  "Performance Data for SW-846 Methods 8270, 8081, and 8141,"
      EMSL-LV, EPA/600/4-90/015.

11.   Marsden, P.O., "Analysis of PCBs", EMSL-LV, EPA/600/8-90/004

12.   Erickson, M. Analytical  Chemistry of PCBs, Butterworth Publishers,  Ann
      Arbor  Science  Book (1986).

13.   Stewart, J.  "EPA Verification Experiment  for Validation of the SOXTEC" PCB
      Extraction Procedure";  Oak Ridge National Laboratory,  Oak Ridge,  TN,
      37831-6138; October 1988.

14.   Lopez-Avila,  V. (Beckert,  W.,  Project Officer),  "Development of a Soxtec
      Extraction Procedure  for Extracting  Organic Compounds  from Soils  and
      Sediments", EPA  600/X-91/140,  US EPA, Environmental Monitoring  Systems
      Laboratory-Las Vegas, October 1991.
                                  8081 - 23                         Revision 0
                                                                September 1994

-------
15.    Stewart,  J.H.; Bayne, C.K.; Holmes,  R.L.;  Rogers,  W.F.;  and Maskarinec,
      M.P.,  "Evaluation of a Rapid .Quantitative  Organic  Extraction System for
      Determining the Concentration of PCB in Soils", Proceedings of the USEPA
      Symposium  on  Waste Testing  and Quality Assurance.  Oak  Ridge  National
      Laboratory, Oak Ridge,  TN 37831-6131; July  11-15,  1988.
                                   8081  -  24                        Revision 0
                                                                September 1994

-------
                           TABLE 1

  GAS CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION
LIMITS FOR THE ORGANOCHLORINE PESTICIDES AND PCBs AS AROCLORS
              USING WIDE-BORE CAPILLARY  COLUMNS
              SINGLE COLUMN METHOD OF ANALYSIS

Compound
Aldrin
a-BHC
6-BHC
£-BHC
y-BHC (Lindane)
a-Chlordane
7-Chlordane
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dieldrin
Endosulfan I
Endosulfan II
Endosulfan Sulfate
Endrin
Endrin aldehyde
Heptachlor
Heptachlor epoxide
Methoxychlor
Toxaphene
Aroclor-1016
Aroclor-1221
Aroclor-1232
Aroclor-1242
Aroclor-1248
Aroclor-1254
Aroclor-1260
Retention
DB 608b
11.84
8.14
9.86
11.20
9.52
15.24
14.63
18.43
16.34
19.48
16.41
15.25
18.45
20.21
17.80
19.72
10.66
13.97
22.80
MR
MR
MR
MR
MR
MR
MR
MR
Water = Organic-free reagent
Time (min)
DB 1701b
12.50
9.46
13.58
14.39
10.84
16.48
16.20
19.56
16.76
20.10
17.32
15.96
19.72
22.36
18.06
21.18
11.56
15.03
22.34
MR
MR
MR
MR
MR
MR
MR
MR
water.
MDLa Water
(M9/L)
0.034
0.035
0.023
0.024
0.025
0.008
0.037
0.050
0.058
0.081
0.044
0.030
0.040
0.035
0.039
0.050
0.040
0.032
0.086
NA
0.054
NA
NA
NA
NA
NA
0.90

MDLa Soil
(M9/kg)
2.2
1.9
3.3
1.1
2.0

1.5
4.2
2.5
3.6
NA
2.1
2.4
3.6
3.6
1.6
2.0
2.1
5.7
NA
57.0
NA
NA
NA
NA
NA
70.0

Soil = Sandy loam soil .
MR = Multiple
NA = Data not
peak responses.
available.



MDL  is  the method detection  limit.   MDL was  determined from the
analysis  of  seven  replicate  aliquots  of  each matrix  processed
through  the  entire  analytical   method   (extraction,  silica  gel
cleanup, and GC/ECD analysis).  MDL = t(n-l,  0.99) x SD, where  t(n-
1, 0.99) is the Student's t value  appropriate for a 99%  confidence
interval and  a standard deviation with n-1 degrees of freedom, and
SD is the standard deviation of the seven replicate measurements.
See Table 4 for GC operating conditions.
                       8081  -  25
    Revision 0
September 1994

-------
                                 TABLE  2

        GAS CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION
      LIMITS  FOR THE  ORGANOCHLORINE  PESTICIDES  AND  PCBs AS AROCLORS
                   USING  NARROW-BORE CAPILLARY  COLUMNS
                    SINGLE COLUMN METHOD OF ANALYSIS
Compound
Retention Time (min)
  DB 608b    DB 5b
MDLa Water MDL8 Soil
 (M9/L)     (M9/kg)
Aldrin
<7-BHC
6-BHC

-------
                           TABLE  2
                         (Continued)

MDL  is  the  method detection  limit.   MDL was  determined  from the
analysis  of seven  replicate  aliquots  of  each matrix  processed
through  the entire  analytical method  (extraction,  cleanup,  and
GC/ECD analysis).  MDL = t(n-l, 0.99) x SD, where  t(n-l,  0.99) is
the Student's t value appropriate for  a 99% confidence interval and
a standard  deviation with n-1 degrees of freedom, and SD  is the
standard deviation of the seven replicate measurements.

30 m x  0.25 mm ID DB-608 1 /urn film thickness,  see Table  4  for GC
operating conditions.

30 m x  0.25 mm  ID DB-5  1 jum  film  thickness,  see Table  4  for GC
operating conditions.
                      8081 - 27                         Revision 0
                                                    September 1994

-------
                                  TABLE 3

DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQLs) FOR VARIOUS MATRICES8



    Matrix                                                   Factor
    Ground water                                                  10
    Low-concentration soil by sonication with GPC cleanup        670
    High-concentration soil and sludges by sonication         10,000
    Non-water miscible waste                                 100,000
    EQL = [Method detection limit for water (see Table 1 or Table 2) wide-
    bore or narrow-bore options] x [Factor found in this table].  For
    nonaqueous samples, the factor is on a wet-weight basis.  Sample EQLs
    are highly matrix-dependent.  The EQLs to be determined herein are
    provided for guidance and may not always be achievable.
                                 8081 - 28                         Revision 0
                                                              September  1994

-------
                                   TABLE 4
             GC OPERATING CONDITIONS FOR ORGANOCHLORINE COMPOUNDS
                            SINGLE  COLUMN ANALYSIS
      Narrow-bore  columns:

Narrow-bore Column 1 - 30 m x 0.25 or 0.32 mm  internal diameter  (ID)  fused
silica capillary column chemically bonded with SE-54  (DB-5  or  equivalent), 1
/im film thickness.
      Carrier  gas  (He)
      Injector temperature
      Detector temperature
      Initial  temperature
      Temperature  program

      Final  temperature
16 psi
225°C
300°C
100°C, hold 2 minutes
100°C to 160°C at 15°C/min,  followed
by 160°C to 270°C at 5°C/nnr)
270°C
Narrow-bore Column 2 - 30 m x 0.25 mm  ID fused silica capillary  column
chemically bonded with 35 percent phenyl methylpolysiloxane  (DB-608,  SPB-608,
or equivalent), 25 jitm coating thickness, 1  /urn film thickness
      Carrier gas  (N2)
      Injector temperature
      Detector temperature
      Initial temperature
      Temperature  program
      Final temperature
20 psi
225°C
300°C
160°C,
160°C
290°C,
 hold 2 minutes
to 290°C at 5°C/min
 hold 1 min
      Wide-bore columns:

Wide-bore Column 1 - 30 m x 0.53 mm ID fused silica capillary column
chemically bonded with 35 percent phenyl methylpolysiloxane  (DB-608,  SPB-608,
RTx-35,  or equivalent), 0.5 /j,m  or 0.83 ,um  film thickness.

Wide-bore Column 2 - 30 m x 0.53 mm ID fused silica capillary column
chemically bonded with 50 percent phenyl methylpolysiloxane  (DB-1701, or
equivalent), 1.0 urn film thickness.
      Carrier gas  (He)
      Makeup gas
      argon/methane  (P-5 or  P-10) or N2
      Injector temperature
      Detector temperature
      Initial temperature
      Temperature  program
      Final temperature
5-7 mL/minute

30 mL/min
250°C
290°C
150°C,  hold 0.5 minute
150°C  to  270°C  at  5°C/min
270°C,  hold 10  min
                                                                    (continued)
                                   8081  -  29
                          Revision 0
                      September 1994

-------
                             TABLE 4   (Continued)
             GC  OPERATING CONDITIONS  FOR ORGANOCHLORINE COMPOUNDS
                            SINGLE COLUMN ANALYSIS
Wide-bore Columns (continued)

Wide-bore Column 3 - 30 m x 0.53 mm ID fused silica capillary column
chemically bonded with SE-54 (DB-5, SPB-5, RTx-5, or equivalent),  1.5
thickness.

      Carrier gas (He)                    6 mL/minute
      Makeup gas
      argon/methane  (P-5 or P-10)  or  N2   30 mL/min
      Injector temperature                205°C
      Detector temperature                290°C
      Initial temperature                 140°C,  hold  2  min
      Temperature program                 140°C  to  240°C at  10°C/min,
                                          hold 5 minutes at 240°C,
                                          240°C  to  265°C at  5°C/min
      Final temperature                   265°C,  hold  18 min
                                   8081 - 30                         Revision 0
                                                                 September 1994

-------
                    TABLE 5
RETENTION TIMES OF THE ORGANOCHLORINE PESTICIDES'
         DUAL COLUMN METHOD  OF ANALYSIS
Compound
DBCP
Hexachl orocycl opentadi ene
Etridiazole
Chloroneb
Hexachl orobenzene
Dial late
Propachlor
Trifluralin
a-BHC
PCNB
7-BHC
Heptachlor
Aldrin
Alachlor
Chlorothalonil
Alachlor
/3-BHC
Isodrin
DC PA
5-BHC
Heptachlor epoxide
Endosulfan-I
•y-Chlordane
a-Chlordane
trans-Nonachlor
4,4'-DDE
Dieldrin
Captan
Perthane
Endrin
Chloropropylate
Chi orobenzi late
Nitrofen
4,4'-DDD
Endosulfan II
4,4'-DDT
Endrin aldehyde
Mi rex
Endosulfan sulfate
CAS No.
96-12-8
77-47-4
2593-15-9
2675-77-6
118-74-1
2303-16-4
1918-16-17
1582-09-8
319-84-6
82-68-8
58-89-9
76-44-8
309-00-2
15972-60-8
1897-45-6
15972-60-8
319-85-7
465-73-6
1861-32-1
319-86-8
1024-57-3
959-98-8
5103-74-2
5103-71-9
39765-80-5
72-55-9
60-57-1
133-06-2
72-56-0
72-20-8
99516-95-7
510-15-6
1836-75-5
72-54-8
33213-65-9
50-29-3
7421-93-4
2385-85-5
1031-07-8
DB-5
RT(min)
2.14
4.49
6.38
7.46
12.79
12.35
9.96
11.87
12.35
14.47
14.14
18.34
20.37
18.58
15.81
18.58
13.80
22.08
21.38
15.49
22.83
25.00
24.29
25.25
25.58
26.80
26.60
23.29
28.45
27.86
28.92
28.92
27.86
29.32
28.45
31.62
29.63
37.15
31.62
DB-1701
RT(min)
2.84
4.88
8.42
10.60
14.58
15.07
15.43
16.26
17.42
18.20
20.00
21.16
22.78
24.18
24.42
24.18
25.04
25.29
26.11
26.37
27.31
28.88
29.32
29.82
30.01
30.40
31.20
31.47
32.18
32.44
34.14
34.42
34.42
35.32
35.51
36.30
38.08
38.79
40.05
                                                     continued
                   8081 -  31
    Revision 0
September 1994

-------



Compound
Methoxychlor
Captafol
Endrin ketone
trans-Permethrin
Kepone
Dicofol
Dichlone
a,a'-Dibromo-m-xylene
2-Bromobiphenyl
TABLE 5
(Continued)

CAS No.
72-43-5
2425-06-1
53494-70-5
51877-74-8
143-50-0
115-32-2
117-80-6




DB-5
RT(min)
35.33
32.65
33.79
41.50
31.10
35.33
15.17
9.17
8.54


DB-1701
RT(min)
40.31
41.42
42.26
45.81
b
b
b
11.51
12.49
3The GC operating conditions were as follows:  30-m x 0.53-mm ID DB-5
 (0.83-jum film thickness) and 30-m x 0.53-mm  ID DB-1701  (1.0-/im film
 thickness)  connected to an 8-in injection tee (Supelco  Inc.).  Temperature
 program: 140°C  (2-min hold)  to  270°C  (1-min  hold)  at   2.8°C/min; injector
 temperature 250°C;  detector temperature 320°C; helium carrier gas  6 mL/min;
 nitrogen makeup gas 20 mL/min.
     detected at 2 ng per injection.
                                   8081 - 32                         Revision  0
                                                                 September 1994

-------
Column 1:
                                    TABLE 6
              GC OPERATING CONDITIONS FOR ORGANOCHLORINE PESTICIDES
                       FOR DUAL COLUMN METHOD OF ANALYSIS
                          LOW TEMPERATURE, THIN FILM
            Type:  DB-1701  (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm  ID
            Film Thickness  (jiim):   1.0
Column 2:
            Type:  DB-5  (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm  ID
            Film Thickness (fj,m):   0.83

Carrier gas flowrate  (mL/min):  6  (Helium)

Makeup gas flowrate (mL/min):  20  (Nitrogen)

Temperature program:  140°C (2 min hold) to 270°C (1  min hold) at 2.8°C/min

Injector temperature:  250°C

Detector temperature:  320°C

Injection volume:  2 p.1

Solvent:  Hexane

Type of injector:  Flash vaporization

Detector type:  Dual ECD

Range:  10

Attenuation:  64 (DB-1701J/32 (DB-5)

Type of splitter:  Supelco 8 in injection tee
                                   8081  -  33                         Revision 0
                                                                September 1994

-------
Column 1:
                                   TABLE 7
             GC OPERATING CONDITIONS FOR ORGANOCHLORINE PESTICIDES
                    FOR THE DUAL COLUMN METHOD OF ANALYSIS
                         HIGH TEMPERATURE, THICK FILM
            Type:  DB-1701 (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm ID
            Film Thickness:  1.0 p,m
Column 2:
            Type:  DB-5 (J&W) or equivalent
            Dimensions:  30 m x 0.53 mm ID
            Film Thickness:  1.5 urn
Carrier gas flowrate (mL/min):  6 (Helium)
Makeup gas flowrate (mL/min):  20 (Nitrogen)
Temperature program:  150°C (0.5 min  hold) to 190°C  (2  min  hold)  at  12°C/min
                      then to 275°C  (10 min hold)  at 4°C/min.
Injector temperature:  250°C
Detector temperature:  320°C
Injection volume:  2 til
Solvent:  Hexane
Type of injector:  Flash vaporization
Detector type:  Dual ECD
Range:  10
Attenuation:  64 (DB-170IJ/64 (DB-5)
Type of splitter:  J&W Scientific press-fit Y-shaped inlet splitter
                                   8081  - 34                        Revision 0
                                                                September  1994

-------
              TABLE 8   SUMMARY OF  RETENTION TIMES (MIN) OF  AROCLORS
                                ON  THE DB-5 COLUMN8
                              DUAL  SYSTEM OF ANALYSIS
Peak
No.b
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41

42
43
44
45
46
47
48

49
50
51
52
53
54
55
Aroclor
1016


8.41
8.77
8.98
9.71
10.49
10.58
10.90
11.23
11.88
11.99
12.27
12.66
12.98
13.18
13.61
13.80
13.96
14.48
14.63
14.99
15.35
16.01

16.27































Aroclor
1221
5.85
7.63
8.43
8.77
8.99

10.50
10.59

11.24


12.29
12.68
12.99










































Aroclor
1232
5.85
7.64
8.43
8.78
9.00

10.50
10.59
10.91
11.24
11.90
12.00
12.29
12.69
13.00
13.19
13.63
13.82
13.97
14.50
14.64
15.02
15.36

16.14
16.29

17.04
17.22
17.46



18.41
18.58

18.83
19.33


20.03





21.18










Aroclor
1242

7.57
8.37
8.73
8.94
9.66
10.44
10.53
10.86
11.18
11.84
11.95
12.24
12.64
12.95
13.14
13.58
13.77
13.93
14.46
14.60
14.98
15.32
15.96
16.08
16.26


17.19
17.43

17.92
18.16
18.37
18.56

18.80
19.30


19.97


20.46

20.85
21.14



22.08






Aroclor
1248




8.95

10.45

10.85
11.18
11.85

12.24
12.64
12.95
13.15
13.58
13.77
13.93
14.45
14.60
14.97
15.31

16.08
16.24

16.99
17.19
17.43
17.69
17.91
18.14
18.36
18.55

18.78
19.29


19.92


20.45

20.83
21.12
21.36


22.05






Aroclor
1254
















13.59
13.78
13.90
14.46

14.98
15.32

16.10
16.25
16.53
16.96
17.19
17.44
17.69
17.91
18.14
18.36
18.55

18.78
19.29
19.48
19.81
19.92

20.28

20.57
20.83
20.98
21.38
21.78

22.04
22.38
22.74
22.96
23.23

23.75
Aroclor
1260
















13.59








16.26

16.97
17.21




18.37

18.68
18.79
19.29
19.48
19.80


20.28

20.57
20.83

21.38
21.78

22.03
22.37
22.73
22.95
23.23
23.42
23.73
Pesticide elating at same
retention time









Chlorothalonil (11.18)















Captan (16.21)

gamma-Chlordane (16.95)





4,4'-DDE (18.38)
Dieldrin (18.59)





Chloropropylate (19.91)
Endosulfan II (19.91)




Kepone (20.99)

4,4'-DDT (21.75)
Endosulfan sulfate (21.75)


Captafol (22.71)



Endrin ketone (23.73)
"The GC operating conditions are given in Table 7.
                                                                            (continued)
                                     8081  - 35
    Revision 0
September 1994

-------
                                     TABLE 8   CONTINUED
Peak
No.
56
57
58

59
60
61
62
63
64
65
66
67
68
69
Aroclor Aroclor Aroclor Aroctor Aroclor Aroclor
1016 1221 1232 1242 1248 1254
23.99

24.27


24.61
24.93

26.22






Aroclor Pesticide e luting at same
1260 retention time
23.97
24.16
Methoxychlor (24.29)
Dicofol (24.29)
24.45
24.62
24.91
25.44
26.19 Mirex (26.19)
26.52
26.75
27.41
28.07
28.35
29.00
"The GC operating conditions are given in Table 7.
'"These are sequentially numbered from elution order and are not isomer numbers
                                           8081  - 36
     Revision  0
September 1994

-------
              TABLE 9   SUMMARY OF  RETENTION TIMES  (MIN) OF AROCLORS
                               ON THE DB-1701 COLUMN8
                              DUAL  SYSTEM OF ANALYSIS
Peak
No.b
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
Aroclor
1016




6.33
6.78
6.96
7.64
8.23
8.62
8.88
9.05
9.46
9.77
10.27
10.64

11.01
11.09
11.98
12.39

12.92
12.99
13.14

13.49
13.58


























Aroclor
1221
4.45
5.38
5.78
5.86
6.34
6.78
6.96

8.23
8.63

9.06

9.79
10.29
10.65






































Aroctor
1232
4.45


5.86
6.34
6.79
6.96

8.23
8.63
8.89
9.06
9.47
9.78
10.29
10.66

11.02
11.10
11.99
12.39
12.77

13.00
13.16

13.49
13.61

14.08
14.30

14.49


15.38
15.65
15.78
16.13




16.77
17.13









Aroclor
1242




6.28
6.72
6.90
7.59
8.15
8.57
8.83
8.99
9.40
9.71
10.21
10.59
10.96
11.02

11.94
12.33
12.71
12.94
13.09


13.44
13.54
13.67
14.03
14.26

14.46


15.33
15.62
15.74
16.10




16.73
17.09

17.46
17.69

18.48


19.13

Aroclor
1248






6.91

8.16

8.83
8.99
9.41
9.71
10.21
10.59
10.95
11.03

11.93
12.33
12.69
12.93
13.09


13.44
13.54

14.03
14.24
14.39
14.46

15.10
15.32
15.62
15.74
16.10




16.74
17.07

17.44
17.69
18.19
18.49


19.13

Aroclor
1254
















10.95


11.93
12.33


13.10

13.24

13.51
13.68
14.03
14.24
14.36

14.56
15.10
15.32
15.61
15.74
16.08

16.34
16.44
16.55
16.77
17.07
17.29
17.43
17.68
18.17
18.42
18.59
18.86
19.10
19.42
Aroclor Pesticide e luting at same
1260 retention time






Trifluralin (6.96)




















13.52

14.02
14.25


14.56

Chlordane (15.32)
16.61 4,4'-DDE (15.67)
15.79

16.19
16.34
16.45

16.77 Perthane (16.71)
17.08
17.31
17.43
17.68
18.18
18.40

18.86
19.09 Endosulfan II (19.05)
19.43
"The GC operating conditions are given in Table 7.
                                                                            (continued)
                                     8081  - 37
    Revision 0
September 1994

-------
                                     TABLE 9   CONTINUED
Peak
No.
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
Aroclor Aroclor Aroclor Aroclor Aroclor Aroclor
1016 1221 1232 1242 1248 1254
19.55
20.20
20.34

20.57 20.55
20.62
20.88

21.53
21.83
23.31








Aroclor Pesticide e luting at same
1260 retention time
19.59 4,4'-DDT (19.54)
20.21

20.43

20.66 Endrin aldehyde (20.69)
20.87
21.03
21.53
21.81
23.27
23.85
24.11
24.46
24.59
24.87
25.85
27.05
27.72
'The GC operating conditions are given in Table 7.
""These are sequentially numbered from elution order and are not isomer numbers
                                           8081  - 38
     Revision  0
September 1994

-------
                                   TABLE 10
            PEAKS DIAGNOSTIC OF PCBs OBSERVED IN 0.53 mm ID COLUMN
                            SINGLE COLUMN ANALYSIS
Peak         RT on      RT on                           Elution
 No.c       DB 608°   DB 1701a Aroclorb                  Order

I479047661221Before TCmX

II           7.15        6.96  1221, 1232, 1248         Before a-BHC

III          7.89        7.65  1061, 1221, 1232, 1242,  Before a-BHC

IV           9.38        9.00  1016, 1232, 1242, 1248,  just after a-BHC on
                                                        DB-1701;just before
                                                        7-BHC  on DB-608

V           10.69       10.54  1016, 1232, 1242.        1248 a-BHC  and
                                                        heptachlor on DB-1701;
                                                        just after heptachlor
                                                        on  DB-608

VI          14.24       14.12  1248, 1254               T-BHC  and heptachlor
                                                        epoxide  on DB-1701;
                                                        heptachlor epoxide  and
                                                        7-Chlordane  on DB-608

VII         14.81       14.77  1254                     Heptachlor epoxide  and
                                                        7-Chlordane  on
                                                        DB-1701;  a-  and
                                                        7-Chlordane  on DB-608

VIII        16.71       16.38  1254                     DDE and  Dieldrin on
                                                        DB-1701;  Dieldrin and
                                                        Endrin on DB-608

IX          19.27       18.95  1254, 1260               Endosulfan II on
                                                        DB-1701;  DDT on DB-608
                                                       Continued
                                  8081  - 39                         Revision 0
                                                                September 1994

-------
                             TABLE 10 (Continued)
            PEAKS DIAGNOSTIC OF PCBs OBSERVED IN 0.53 mm ID COLUMN
                            SINGLE  COLUMN  ANALYSIS
Peak         RT on      RT  on                            Elution
 No.        DB 608a   DB 1701"    Aroclorb              Order
X           21.22       21.23       1260                 Endrin  aldehyde and
                                                        Endosulfan  sulfate on
                                                        DB-1701;  Endosulfan
                                                        sulfate and
                                                        Methoxychlor on
                                                        on  DB-608

XI          22.89       22.46       1260                 Just  before endrin
                                                        ketone  on DB-1701;
                                                        after endrin ketone  on
                                                        DB-608
a  Temperature program:  T( =  150°C, hold 30 seconds; increase temperature at
   5°C/minutes to 275°C.

b  Underlined Aroclor indicates the largest peak in the pattern.

0  These are sequentially numbered from elution order and are not isomer
   numbers
                                   8081 - 40                         Revision 0
                                                                September  1994

-------
                 TABLE 11  SPECIFIC PCB CONGENERS IN AROCLORS
Congener
IUPAC number
                  Aroclor
1016  1221   1232  1242  1248  1254  1260
Biphenyl
2CB
23DCB
34DCB
244'TCB
22'35'TCB
23'44'TCB
233'4'6PCB
23'44'5PCB
22'44'55'HCB
22'344'5'HCB
22'344'55'HpCB
22'33'44'5HpCB
..
1
5
12
28*
44
66*
110
118*
153
138
180
170
X
XXX
XXX
X X
X X
X








X
X
X
X
X









X
X
X
X
X










X
X
X
X
X










X

X
X
X
X
X
*apparent co-elution of two major peaks:

      28 with 31 (2,4',5 trichloro)
      66 with 95 (2,2',3,5',6 pentachloro)
      118 with 149 (2,2',3,4',5',6 hexachloro)
                                  8081  - 41
                                                  Revision  0
                                              September  1994

-------
                 TABLE  12   ANALYTE  RECOVERY  FROM  SEWAGE  SLUDGE
Compound
Sonication
Soxhlet

Hexachloroethane
2-Chloronapthalene
4-Bromodiphenyl ether
a-BHC
7-BHC
Heptachlor
Aldrin
j8-BHC
(J-BHC
Heptachlor epoxide
Endosulfan I
7-Chlordane
a-Chlordane
DDE
Dieldrin
Endrin
Endosulfan II
DDT
Endrin aldehyde
ODD
Tetrachl oro-m-xyl ene
Decachl orobi phenyl
%Recovery
80
50
118
88
55
60
92
351
51
54
52
50
49
52
89
56
52
57
45
57
71
26
%RSD
7
56
14
25
9
13
33
71
11
11
11
9
8
11
19
10
10
10
6
11
19
23
%Recovery
79
67
nd
265
155
469
875
150
57
70
70
65
66
74
327
92
88
95
42
99
82
28
%RSD
1
8

18
29
294
734
260
2
3
4
1
0
1
7
15
11
17
10
8
1
48
Concentration spiked in the sample:  500-1000 ng/g
Three replicates/sample

Extraction solvent, Method 3540 - methylene chloride
Extraction solvent, Method 3550 - methylene chloride/acetone (1:1)

Cleanup - Method 3640

GC column - DB-608, 30M X 0.53 mm ID
                                   8081  -  42
                                   Revision 0
                               September 1994

-------
               TABLE  13   ANALYTE  RECOVERY  FROM  DCE  STILL  BOTTOMS
Compound
Sonication
Soxhlet

Hexachloroethane
2-Chloronapthalene
4-Bromodiphenyl ether
a-BHC
/3-BHC
Heptachlor
Aldrin
l8-BHC

-------
                                   TABLE 14
             SINGLE  LABORATORY  ACCURACY DATA FOR THE EXTRACTION OF
        ORGANOCHLORINE PESTICIDES FROM SPIKED CLAY SOIL BY METHOD 3541
                             (AUTOMATED SOXHLET)"
Compound Name                 Spike Level       	% Recovery	
                                                      DB-5              DB-1701
a-BHC
/3-BHC
Heptachlor
Aldrin
Heptachlor epoxide
trans-Chlordane
Endosulfan I
Dieldrin
Endrin
Endosulfan II
4,4'-DDT
Mi rex
500
500
500
500
500
500
500
500
500
500
500
500
89
86
94
b
97
94
92
b
111
104
b
108
94
b
95
92
97
95
92
113
104
104
b
102
a     The  operating conditions  for the  automated  Soxhlet  were  as follows:
      immersion time 45 min;  extraction time  45 min; the sample size was 10 g
      clay soil, extraction solvent, 1:1 acetone/hexane.  No equilibration time
      following spiking.

b     Not able to determine because of interference.

Data taken from Reference  14.
                                   8081 - 44                        Revision 0
                                                                September  1994

-------
                         TABLE 15
     SINGLE LABORATORY RECOVERY DATA FOR EXTRACTION OF
PCBS FROM CLAY AND SOIL BY METHOD 3541a  (AUTOMATED  SOXHLET)
Matrix Compound Spike Level
(ppm)
Clay Aroclor-1254 5





Clay Aroclor-1254 50





Clay Aroclor-1260 5





Clay Aroclor-1260 50





Soil Aroclor-1254 5




Soil Aroclor-1254 50





Trial
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
1
2
3
4
5
6
Percent
Recovery6
87.0
92.7
93.8
98.6
79.4
28.3
65.3
72.6
97.2
79.6
49.8
59.1
87.3
74.6
60.8
93.8
96.9
113.1
73.5
70.1
92.4
88.9
90.2
67.3
69.7
89.1
91.8
83.2
62.5
84.0
77.5
91.8
66.5
82.3
61.6
                                            (continued)
                        8081  -  45
    Revision 0
September 1994

-------
                                   TABLE 15
                                  (continued)
Matrix Compound Spike Level
(ppm)
Soil Aroclor-1260 5






Soil Aroclor-1260 50





Trial
1
2
3
4
5
6
7
1
2
3
4
5
6
Percent
Recovery15
83.9
82.8
81.6
96.2
93.7
93.8
97.5
76.9
69.4
92.6
81.6
83.1
76.0
a     The  operating  conditions  for the  automated Soxhlet  were as  follows:
      immersion time 60 min; reflux time 60 min.

b     Multiple results from two different extractors.

Data from Reference 15.
                                   8081  -  46
    Revision 0
September 1994

-------
            TABLE 16.   MULTI-LABORATORY PRECISION AND ACCURACY DATA
                  FOR  THE EXTRACTION OF PCBS  FROM SPIKED SOIL
                      BY METHOD 3541 (AUTOMATED SOXHLET)
Laboratory
Lab 1
Lab 2
Lab 3
Lab 4
Lab 5
Lab 6
Lab 7
Lab 8
All
Laboratories
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
	
Num
Average
St Dev
Num
Average
St Dev
Num
Average
St Dev
PCB Percent Recovery
Aroclor
1254
PCB Level
5
3.0
101.2
34.9

3.0
72.8
10.8
6.0
112.6
18.2

2 0
i4o!g
4.3
3.0
100.1
17.9
	
3.0
65.0
16.0
20.0
98.8
28.7
	
50
3.0
74.0
41.8
	
6.0
56.5
7.0
3.0
63.3
8.3
	
6.0
144.3
30.4
3.0
97.1
8.7
3.0
127.7
15.5
	
3.0
123.4
14.6
	
3.0
38.3
21.9
30.0
92.5
42.9
500

6.0
66.9
15.4


3.0
80.1
5.1



9.0
71.3
14.1
1260
PCB Level
5
3.0
83.9
7.4

3.0
70.6
2.5
6.0
100.3
13.3

3.0
138.7
15.5
3.0
82.1
7.9
3.0
92.8
36.5
21.0
95.5
25.3
50
3.0
78.5
7.4
6.0
70.1
14.5
3.0
57.2
5.6
6.0
84.8
3.8
3.0
79.5
3.1
4.0
105.9
7.9
3.0
94.1
5.2
3.0
51.9
12.8
	
31.0
78.6
18.0
500

6.0
74.5
10.3


3.0
77.0
9.4



	
9.0
75.3
9.5
All
Levels
12.0
84.4
26.0
24.0
67.0
13.3
12.0
66.0
9.1
24.0
110.5
28.5
12.0
83.5
10.3
12.0
125.4
18.4
12.0
99.9
19.0
12.0
62.0
29.1
120.0
87.6
29.7
Data from Reference 13.
                                   8081  -  47
    Revision 0
September 1994

-------
                                      FIGURE 1.
         GAS CHROMATOGRAM OF  THE MIXED  ORGANOCHLORINE PESTICIDE STANDARD
Start Time : 0.00 min      End Time  : 33.00 *?n
Scale Factor:   0         Plot Offset: 20 mV
                                              lo« Point : 20.00
                                              Plot Scilt: 400 M
                                                                High Point : 420.00 «V
                                       Response  [mV]
       o~
    iD
    78-
                                              NJ
                                                           GJ
                                             ill i 1  1 1  1 1 1  1 1 1  1 1  1 1 1  1 1 1  I
                                                                   LL
                      •= -4.68
                      :-7.99
                          L.67
                                             -9.93
                                -10.78
                                        	11.05
                                     -11.81
                                         -13.65
                         -14.34
                        4l§gS
                                         -14.92
                                     -16.32
                                     	17.17
                                                   -17.63
                                                an—18.56
                                                           -23.18
                             -23.80
                              •26.23
                             -=^28.64
Column:
Temperature program:
                                                                             -0.95
                                                                             -8.60
                                                                         -30.19
                   30 m x 0.25 mm ID,  DB-5
                   100°C  (hold 2  minutes)  to  160°C at  15°C/min,  then  at
                   5°C/min to  270°C;  carrier  He at 16 psi.
                                      8081  -  48
                                                                   Revision  0
                                                               September  1994

-------
                                      FIGURE 2.
     GAS CHROMATOGRAM OF  INDIVIDUAL ORGANOCHLORINE  PESTICIDE STANDARD MIX A
          Start Time : 0.00 mm      End lime  : 33.00 nun
          Scale Factor:  0         Plot Offset: 20 mv
                      lav Point : 20.00 mV      Kign Point : 270.00 mv
                      Plot Scale: 250 mv
                                         Response  [mV]
                        i  i  i     i  i  i
                                         T,  .   ,  ,?
                                    N)
                                    O
                                    O
                                       I  I  I  I  I   I  I
         o"
      m
      no"
      D
      r-t-
      o'
      3

      H
         ro
          -
         LJ
         O"
                         '-7.93
                            1.60
 -14.27
                            -17.08
                                20.22
                                0.77
                                  22.68
                                 -23.73
                                  --28.52
                                                                              -8.54
                                                        •-9.86
-12
.33
                                                      10.98
                           13.58
                                                                   -17.54
                                                                     18.47
                      -19.78
                                                            -19.24
                                                       -21.13
                                                                              -23.03
                                              -30.05
Column:
Temperature program:
30 m x  0.25 mm ID, DB-5
100°C  (hold 2  minutes)  to  160°C  at  15°C/min,  then  at
5°C/min to  270°C;  carrier He at 16 psi.
                                     8081  -  49
                                               Revision 0
                                           September 1994

-------
                                       FIGURE  3
     GAS CHROMATOGRAM OF  INDIVIDUAL  ORGANOCHLORINE PESTICIDE STANDARD MIX  B
        Start rime  : 0.00 mm
        Scale Factor:  0
                       End Time  :  33.00 mm
                       Plot Offset:  20 mv
Low Point : 20.00 mv
Plot Scale: 250 mv
                                                                High Point • 270 20 mv
                                       Response [mV]
                            o
                        i  i  I   i  i
                                   a          01
                                   o          a
                                   I   i  i  i  i  I
              K)         fO
              o         cn
              o         a
        i  i   i  I  i  i  i   i  I  i  i
   Q'


fl>
    H
    i'
    fD
       Ul
       O"
                         2.74

                         -6.97
                       4—9.60
                                   --10.71
                         -14.27
                        ^-15.24
                        -<-i g nf.
                              22.00
                                             -11.73
                                                 -14.84
                                            -16.23
                                          —17.08
                                          -17.63
                                                  -18.31

                                                  	19.54
                                            -20.19
                                            	21.03
                                                        --22.68
                                                                             -9.95
                                                                             -8.5-
                                                               -30.04
Column:
Temperature program:
                      30 m x 0.25  mm ID, DB-5
                      100°C  (hold  2 minutes)  to 160°C  at  15°C/min»  then  at
                      5°C/min  to 270°C;  carrier He  at 16 psi.
                                      8081  -  50
                                                                       Revision  0
                                                                  September  1994

-------
                                     FIGURE 4.
                   GAS CHROMATOGRAM OF THE TOXAPHENE STANDARD


    Star; Tim : 0.00 mm      End Tine : 33.00 nin      Low Point : 20.00 mv      High Point : 80.00
    Scale Factor:   0         Plot Offset: 20 mV         Plot Sealt: 60 UN
                                   Response  [mV]
                  I I I I I I I I II I I I I I I I I II ll I I ll I I II I I I I I I I I II I I I I I I I Ml I I I I I I I I II
   01—
 ID
 rti

 it"
 3
T8-
   Uf
                                                                         24.32
Column:
Temperature program:
                        30 m x  0.25 mm ID, DB-5
                        100°C (hold 2  minutes)  to 160°C  at 15°C/min,  then  at
                        5°C/min to  270°C;  carrier He at  16 psi.
                                   8081  -  51
                                                                         Revision 0
                                                                     September 1994

-------
                                       FIGURE 5.
                   GAS CHROMATOGRAM OF  THE AROCLOR-1016 STANDARD

         Start Time  : 0.00 mm      End Time  : 33.00 mm       Low Point : 20.00 mV      High Point : 120.00 mV
         Scale Factor:  0         Plot Offset: 20 mV         Plot Scale: 100 mV
                                        Response  [mV]
                      NJ         4*          0)         DO         O
                      O         O          O         O         O
                       I I I I I I I I I I I I  I I I I I I I I I I I II I I I I I I I I I I I II I I I I I I I I I I I 1
        Uf
                                  -1.81
                                                         -12.95
                                                                             -1.03
Column:
Temperature program:
30 m x  0.25 mm ID DB-5  fused silica capillary.
100°C  (hold 2  minutes)  to  160°C  at  15°C/niin, then  at
5°C/min to  270°C;  carrier He at 16 psi.
                                     8081  - 52
                                                Revision 0
                                           September 1994

-------
                                     FIGURE 6.
              GAS  CHROMATOGRAM  OF  THE TECHNICAL  CHLORDANE  STANDARD
       Start Time : 0.00 mm
       Scale Factor:  0
End Time  : 33.00 mm      Lou Point : 20.00 mv      High Point : 220.00 mv
Plot Offset: 20 mv         Plot Scale: 200 mv
                                      Response  [rnV]
                                          o
                                          o
                                                          J	L
                                           to
                                           o
                                           1
J	I	I
      Lfl—
   I'D
   nT
   o  -*.
   3  <-»
   H

   3'
      o
       -
                                    .59
                                         -12.92
                                              13.60
                                                                           -0.97
Column:
Temperature program:
30  m  x 0.25 mm  ID  DB-5 fused  silica capillary.
100°C  (hold  2  minutes)  to  160°C at  15°C/min,  then at
5°C/min  to  270°C;  carrier  He at 16 psi.
                                     8081  -  53
                                                Revision 0
                                           September 1994

-------
           DB-1701
   Li
          OB-5
      •T
      u-
FIGURE 7.   GC/ECD chromatogram of Toxaphene analyzed on a DB-5/DB-1701  fused-
            silica open-tubular column pair.  The GC operating conditions were
            as follows:  30 m x 0.53 mm ID DB-5 (1.5-/nm film thickness)  and 30
            m x 0.53 mm  ID  DB-1701  (1.0-jum film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet  splitter.  Temperature program:
            150°C  (0.5  min  hold)  to  190°C  (2 min hold) at 12°C/min then  to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  54
    Revision 0
September 1994

-------
                                                         0>
                                                         o-
              DB-1701
      o 
      u) IM
     Ju
Or
r^
•s
              DB-5
             ~—A.
FIGURE 8.   GC/ECD chromatogram  of Strobane analyzed on  a  DB-5/DB-1701  fused-
            silica open-tubular  column pair.   The  GC  operating conditions were
            as follows:  30 m x  0.53 mm  ID  DB-5  (1.5-jum film thickness)  and 30
            m x 0.53 mm  ID  DB-1701 (1.0-/zm film thickness)  connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C  (0.5 min hold)  to 190°C (2 min  hold) at  12°C/nnn then to  275°C
            (10 min hold) at 40C/nnn.
                                   8081 - 55
                                                       Revision 0
                                                   September 1994

-------
       \r
       0
in
rv
rxi
              JL
m

-------
        I
                bill
                                                       DB-1701
                           o-
                           f-0
                            •CO
         A/.
         U.	^
ULL
                                                      DB-5
                              (• 1*1
                              • I-
                              o  •
FIGURE 10.  GC/ECD  chromatogram of  Aroclor  1221  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.  The GC operating conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-,um film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at  12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081  -  57
                                                  Revision 0
                                              September 1994

-------
                                                   DB-1701
                                                   DB-5
FIGURE 11.  GC/ECD  chromatogram of  Aroclor 1232  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating conditions
            were as follows:  30 m x 0.53 mm ID  DB-5 (1.5-jum film  thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-Mm film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  58
    Revision 0
September 1994

-------
          r-
          O-
          
          "'•?:•«  'A
            r*M W T  .1
          \JL_jw-_UJ
                                                          DB-1701
       I-J
LL
                                                         DB-5
FIGURE 12.   GC/ECD chromatogram  of Aroclor  1242 analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular  column  pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-jum film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-^m film thickness)  connected to a J&W
            Scientific press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold)  to  190°C (2  min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                  8081  - 59
                                                          Revision 0
                                                      September 1994

-------
           Ill
           0
           (u
                           in
                           in
              —in
              -in
                         r«
                     ODOU 0)
   tn <



  Jj
                                                               DB-1701
O  m

(X  ft
                                                              DB-S
            ,
                to  if,
                o-  —
                M  in
*  S-'
FIGURE 13.  GC/ECD  chromatogram  of  Aroclor  1248 analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.  The GC operating conditions
            were as follows:  30 m x  0.53 mm ID DB-5 (1.5-/nm film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-pi  film thickness)  connected to a J&W
            Scientific press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold) to  190°C  (2 min  hold)  at 12°C/min  then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081 - 60
                                             Revision  0
                                         September 1994

-------
                                          ID
                                          r-
                                                             DB-1701
           UO
           0-
           •o

         li
                                                             DB-5
           uuj  UiO —
         0 O>l  u bi 1*1
k>
«
r«
FIGURE 14.  GC/ECD  chromatogram of  Aroclor  1254  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.  The GC operating conditions
            were as follows:  30 m  x 0.53 mm ID DB-5 (1.5-/Lim film thickness) and
            30 m x 0.53 mm  ID  DB-1701  (1.0-/im  film thickness) connected  to a J&W
            Scientific press-fit Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min  hold) to 190°C  (2 min  hold)  at 12°C/min  then  to 275°C
            (10 min hold)  at  4°C/min.
                                   8081 - 61
                                      Revision 0
                                  September 1994

-------
                         OB-1701
                         11	" .***,'! ..
                                                   : S
                        DB-5
FIGURE 15.  GC/ECD  chromatogram of  Aroclor  1260  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-jum  film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at  12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081 - 62
    Revision 0
September 1994

-------
                                                         DB-1701
            r- -
            © -
                                                           a>
                                                           o
                                                                   KI in
                                                                   n 9
                                                                   r- m
                                                                   01 r>i
            • o-
           rsi r-
                                                        DB-5
                         fO          K»TT 0-
                         *»    <3 T  0)
                         01  O ® (M  ID
                        1  .  O T ®  O
                              ® •"  "• ^™» «•» ^
                                       .AN-fA.
FIGURE 16.  GC/ECD  chromatogram  of Halowax  1000  analyzed  on  a  DB-5/DB-1701
            fused-silica  open-tubular  column  pair.   The GC operating conditions
            were as follows:  30 m x 0.53 mm ID  DB-5 (1.5-/im film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-/im film thickness) connected to a J&W
            Scientific  press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold)  to  190°C (2 min hold)  at  12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081  -  63
    Revision 0
September 1994

-------
                                                DB-1701
FIGURE 17.  GC/ECD  chromatogram of  Halowax  1001  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column  pair.  The GC  operating conditions
            were as follows:  30 m x 0.53 mm ID  DB-5 (1.5-^m film  thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-jum film thickness) connected to a J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  64
    Revision 0
September 1994

-------
           aryo-
                 •09
                                                           DB-1701
                             tr»
                   «
                   •o
                   *>

                   »
                                                           DB-5
           «  9 f-
  CD     <•«(-»
v>««in  mm  •  in •
»«*•  in M   •   .  .
*v» •»•  n  .  -e  « »
 • *» •   • «o  ra  IN P»
 **ir»  » m
FIGURE 18.  GC/ECD  chromatogram  of Halowax  1099  analyzed  on a  DB-5/DB-1701
            fused-silica  open-tubular  column  pair.   The GC operating conditions
            were as follows:  30 m x 0.53 mm ID  DB-5 (1.5-jum film thickness) and
            30 m x 0.53 mm ID DB-1701 (1.0-/im film thickness) connected to a J&W
            Scientific  press-fit  Y-shaped inlet splitter.  Temperature program:
            150°C (0.5 min hold) to  190°C (2 min hold)  at  12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081  -  65
               Revision  0
           September  1994

-------

                                                             DB-1701
                            0  »
                            Hi  —
                             0
T  9
a o-
I- «
•o r-
                       DB-5
         «
         r1
                             \J
FIGURE 19.  GC/ECD  chromatogram of  Halowax 1013  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-^m film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-Mm film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/irnn then to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  66
                               Revision 0
                           September 1994

-------
                        OB-1701
                                    i
                       DB-5
FIGURE 20.  GC/ECD  chromatogram of  Halowax 1014  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-/zm film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min hold) to  190°C (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.
                                   8081  -  67
    Revision 0
September 1994

-------
                            DB-1701
                                  f  ".  s
                          v » • n* «   M   .
                          M ««t - »     •

                          *» «* It •%   •  -  -
                     t 3
                             DB-5
 i


 •t •  * " *
 •lL*-«» *• - fk
 Elf*'-? • - •»

J^Jili
                                           its
FIGURE 21.   GC/ECD chromatogram  of Halowax  1051  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular  column pair.  The GC operating conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/im film thickness)  and
            30 m x 0.53 mm ID DB-1701 (1.0-/im  film thickness) connected to a  J&W
            Scientific press-fit  Y-shaped  inlet splitter.  Temperature program:
            150°C (0.5 min hold)  to  190°C (2 min  hold)  at 12°C/nnn then to 275°C
            (10 min hold) at 4°C/min.
                                  8081  - 68
                                                Revision 0
                                            September 1994

-------
                                     DB-5

I

i I


i
,
1


aw

I
                                     DB-1701
     12      ]    « SU IS  «
                      JL_
                              10 11 12  IS
                                       14
                                                  n
                                                         32
IS   3*   4
          4)
                                                                             20.
FIGURE 22.  GC/ECD chromatogram of  the  organochlorine pesticides analyzed on a
            DB-5/DB-1701  fused-silica   open-tubular  column  pair.    The  GC
            operating conditions were as follows:  30 m x 0.53 mm ID DB-5 (0.83-
            nm  film thickness)  and 30  m x 0.53  mm  ID  DB-1701  (1.0-/um  film
            thickness)  connected to  an  8 in  injection  tee   (Supelco  Inc.).
            Temperature program:   140°C (2 min hold)  to 270°C  (1  min hold)  at
            2.8°C/min.
                                   8081 - 69
       Revision 0
   September  1994

-------
                                        METHOD  8081

               ORGANOCHLORINE PESTICIDES AND PCBs  AS AROCLORS BY  GAS
                      CHROMATOGRAPHY:  CAPILLARY  COLUMN TECHNIQUE
     7.1.1 Choose
  appropriate extraction
technique (see Chapter 2)
         i
   7.1.2 Add specified
  matrix sp*e to sample.
         1
  7.2 Routine cleanup/
     fracttonaaon.
 7.3 Set chrornalographic
         i
7.4 Refer to Method 8000
  tor proper calfcratton
     techniques.
 7.4.2 Prime or deactivate QC
 column prior to caNbratton.
7.5 Perform GO analysis (see
      MettwdSOOO)
         7.5.8
      Any sample
       peak Inter-
       ferences?
7.5.8 Additional
                                                                       (see Section 7.2)
                                       7.8CataJtetionof
                                  toxaphene, chterdane, PCBs,
                                   DDT. and BHC done here.
                                        8081  -  70
                                               Revision 0
                                          September  1994

-------
                                LIST OF TABLES
Table 1     Gas chromatographic retention times and method detection limits for
            the Organochlorine Pesticides and PCBs as Aroclors using wide-bore
            capillary columns, single column analysis
Table 2     Gas chromatographic retention times and method detection limits for
            the Organochlorine pesticides and PCBs as  Aroclors using narrow-bore
            capillary columns, single column analysis
Table 3     Estimated quantitation limits (EQL) for various matrices
Table 4     GC Operating conditions for Organochlorine compounds, single column
            analysis
Table 5     Retention times of the Organochlorine pesticides, dual column method
            of analysis
Table 6     GC operating conditions for Organochlorine pesticides,  dual column
            method of analysis, low temperature, thin film
Table 7     GC operating conditions for Organochlorine pesticides,  dual column
            method of analysis, high temperature, thick film
Table 8     Summary of  retention  times (min)  of Aroclors on  the DB  5 column,
            dual system of analysis
Table 9     Summary of retention times  (min) of Aroclors on  the DB 1701 column,
            dual system of analysis
Table 10    Peaks  diagnostic  of PCBs  observed in 0.53  mm  ID  column,  single
            column system of analysis
Table 11    Specific Congeners in Aroclors
Table 12    Recovery from Sewage Sludge
Table 13    Recovery DCE still bottoms
Table 14    Single Laboratory Accuracy  Data for the Extraction  of Organochlorine
            Pesticides from Spiked Clay Soil by  Method 3541  (Automated Soxhlet)
Table 15    Single Laboratory Recovery  Data for Extraction of PCBs from Clay and
            Soil by Method 3541 (Automated Soxhlet)
Table 16    Multi-laboratory Precision and Accuracy Data for the Extraction of
            PCBs from Spiked Soil  by Method 3541 (Automated Soxhlet)
                                   8081  -  71
    Revision 0
September 1994

-------
                                LIST OF FIGURES

Figure 1.   GC of the Mixed Organochlorine  Pesticide Standard.  The GC operating
            conditions  were   as  follows:   30  m x  0.25  mm  ID  DB-5  column.
            Temperature program:   100°C  (hold  2  minutes)  to 160°C at 15°C/min,
            then at 5°C/min to 270°C;  carrier He at  16 psi.

Figure 2.   GC of  Individual  Organochlorine Pesticide Standard Mix A.   The  GC
            operating  conditions were  as  follows:    30 m x  0.25  mm  ID  DB-5
            column.   Temperature program:   100°C (hold 2  minutes)  to 160°C  at
            15°C/nrin> then at 5°C/mir\  to 270°C;  carrier  He  at 16  psi.

Figure 3.   GC of  Individual  Organochlorine Pesticide Standard Mix B.   The  GC
            operating  conditions were  as  follows:    30 m x  0.25  mm  ID  DB-5
            column.   Temperature program:   100°C (hold 2  minutes)  to 160°C  at
            15°C/min> then at 5°C/min  to 270°C;  carrier  He  at 16  psi.

Figure 4.   GC of the Toxaphene  Standard.   The GC operating conditions  were  as
            follows:   30 m  x 0.25 mm  ID  DB-5  column.   Temperature program:
            100°C (hold 2 minutes) to 160°C  at 15°C/min, then at 5°C/min to 270°C;
            carrier He  at  16  psi.

Figure 5.   GC of the Aroclor-1016  Standard.  The GC  operating conditions  were
            as follows:  30 m x  0.25 mm  ID DB-5  fused silica capillary  column.
            Temperature program:   100°C  (hold  2  minutes)  to 160°C at 15°C/min,
            then at 5°C/min to 270°C;  carrier He at  16 psi.

Figure 6.   GC of the Technical Chlordane Standard.  The GC operating conditions
            were as  follows:   30 m x 0.25 mm  ID DB-5  fused  silica  capillary
            column.   Temperature program:   100°C (hold 2  minutes)  to 160°C  at
            15°C/niin, then at 5°C/min  to 270°C;  carrier  He  at 16  psi.

Figure 7.   GC/ECD chromatogram  of  Toxaphene analyzed on  a DB-5/DB-1701 fused-
            silica open-tubular  column  pair.  The GC  operating conditions  were
            as follows:  30 m x  0.53 mm  ID DB-5  (1.5-jum film thickness) and  30
            m x  0.53  mm ID DB-1701  (1.0-jum film thickness)  connected to a J&W
            Scientific press-fit Y-shaped  inlet splitter.  Temperature program:
            150°C (0.5 min hold)  to 190°C (2 min  hold) at 12°C/min then to 275°C
            (10 min hold)  at  4°C/min.

Figure 8.   GC/ECD chromatogram  of Strobane analyzed on a  DB-5/DB-1701 fused-
            silica open-tubular  column  pair.  The GC  operating conditions  were
            as follows:  30 m x  0.53 mm  ID DB-5  (1.5-/um film thickness) and  30
            m x  0.53  mm ID DB-1701  (1.0-/um film thickness)  connected to a J&W
            Scientific press-fit Y-shaped  inlet splitter.  Temperature program:
            150°C (0.5 min hold)  to 190°C (2 min  hold) at  12°C/min then to 275°C
            (10 min  hold)  at  4°C/min.
                                   8081 - 72
    Revision 0
September 1994

-------
Figure 9.   GC/ECD  chromatogram of  Aroclor  1016  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/im film thickness)  and
            30 m x 0.53  mm  ID DB-1701 (1.0-/zm film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min.

Figure 10.  GC/ECD  chromatogram of  Aroclor  1221  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/Ltm film thickness)  and
            30 m x 0.53  mm  ID DB-1701 (1.0-/xni film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                                                         i
Figure 11.  GC/ECD  chromatogram of  Aroclor  1232  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/xm film thickness)  and
            30 m x 0.53  mm  ID DB-1701 (1.0-pi film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min.

Figure 12.  GC/ECD  chromatogram of  Aroclor  1242  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-/im film thickness)  and
            30 m x 0.53  mm  ID DB-1701 (1.0-jun film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min.

Figure 13.  GC/ECD  chromatogram of  Aroclor  1248  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-jum film thickness)  and
            30 m x 0.53  mm  ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min-

Figure 14.  GC/ECD  chromatogram of  Aroclor  1254  analyzed  on  a  DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:  30 m x 0.53 mm ID DB-5 (1.5-jum film thickness)  and
            30 m x 0.53  mm  ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold)  at 4°C/min.
                                   8081  -  73
    Revision 0
September 1994

-------
Figure 15.  GC/ECD  chromatogram of  Aroclor 1260  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-|um film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/xm film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 16.  GC/ECD  chromatogram of  Halowax 1000  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-Mm film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/niin then to 275°C
            (10 min hold) at 4°C/nrin.

Figure 17.  GC/ECD  chromatogram of, Halowax 1001  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-Mtn film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 18.  GC/ECD  chromatogram of  Halowax 1099  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-/im film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 19.  GC/ECD  chromatogram of  Halowax 1013  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-/im film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-^tm film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 20.  GC/ECD  chromatogram of  Halowax 1014  analyzed on  a DB-5/DB-1701
            fused-silica open-tubular column pair.   The GC  operating  conditions
            were as follows:   30 m x 0.53 mm ID DB-5 (1.5-jum film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-jum film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature program:
            150°C (0.5 min  hold) to  190°C  (2 min hold)  at 12°C/miri then to 275°C
            (10 min hold) at 4°C/min.
                                   8081 - 74
    Revision 0
September 1994

-------
Figure 21.  GC/ECD  chromatogram of  Halowax 1051  analyzed on  a DB-5/D8-1701
            fused-silica open-tubular column pair.   The  GC  operating conditions
            were as follows:   30 m x 0.53 mm ID  DB-5 (1.5-jttm film thickness)  and
            30 m x 0.53 mm  ID DB-1701 (1.0-^m film thickness) connected to a  J&W
            Scientific press-fit Y-shaped inlet splitter.   Temperature  program:
            150°C (0.5 min  hold) to  190°C (2 min hold) at 12°C/min then to 275°C
            (10 min hold) at 4°C/min.

Figure 22.  GC/ECD chromatogram of the organochlorine pesticides analyzed on a
            DB-5/DB-1701  fused-silica  open-tubular  column  pair.    The  GC
            operating conditions were as follows:   30 m x 0.53 mm ID DB-5  (0.83-
            jum  film thickness) and  30 m  x 0.53  mm  ID DB-1701  (1.0-jum film
            thickness)  connected  to  an 8  in   injection   tee  (Supelco  Inc.).
            Temperature program:  140°C  (2  min hold)  to 270°C (1 min  hold) at
            2.8°C/min.
                                  8081  -  75                         Revision 0
                                                                September 1994

-------
                                 METHOD 8090

                      NITROAROMATICS AND CYCLIC KETONES
1.0  SCOPE AND APPLICATION

     1.1  Method 8090  is  used  to  determine  the  concentration  of various
nitroaromatic and cyclic ketone compounds.    Table 1 indicates compounds that
may be determined by this method and lists the method detection limit for each
compound in reagent water.    Table  2  lists the practical quantisation limit
(PQL) for other matrices.


2.0  SUMMARY OF METHOD

     2.1  Method  8090  provides   gas   chromatographic  conditions  for  the
detection of ppb levels of  nitroaromatic  and cyclic ketone compounds.  Prior
to use of this method, appropriate  sample extraction techniques must be used.
Both neat and diluted  organic  liquids  (Method  3580, Waste Dilution) may be
analyzed by direct injection.  A 2- to 5-uL aliquot of the extract is injected
into a gas chromatograph  (GC) using the solvent flush technique, and compounds
in the GC effluent are  detected  by  an  electron capture detector (ECD) or a
flame ionization detector   (FID).    The  dinitrotoluenes are determined using
ECD, whereas the other compounds amenable  to this method are determined using
FID.

     2.2  If interferences  prevent  proper  detection  of  the  analytes, the
method may also be performed on extracts that have undergone cleanup.


3.0  INTERFERENCES

     3.1  Refer to Method 3500, 3600, and 8000.

     3.2  Solvents,  reagents, glassware,  and other  sample-processing  hardware
may   yield   discrete    artifacts    and/or   elevated   baselines    causing
misinterpretation of gas   chromatograms.    All  of  these  materials  must be
demonstrated to  be   free   from  interferences,  under the  conditions of the
analysis, by analyzing   method  blanks.    Specific   selection  of  reagents and
purification of solvents by distillation  in  all-glass systems may  be  required.

     3.3   Interferences  coextracted  from  samples  will vary considerably from
source  to  source, depending upon   the  waste being  sampled.  Although general
cleanup techniques are  recommended as part   of  this  method, unique samples may
require additional cleanup.
                                   8090  -  1
                                                          Revision       0
                                                          Date  September  1986

-------
TABLE 1.  GAS CHROMATOGRAPHY OF NITROAROMATICS AND ISOPHORONE

                                Retention time (m1n)      Method detection
                                                            limit (ug/L)
Compound                         Col. la    Col. 2b        ECD        FID
Isophorone
Nitrobenzene
2,4-D1n1trotoluene
2,6-D1n1trotoluene
Di nitrobenzene
Naphthoqulnone
4.49
3.31
5.35
3.52


5.72
4.31
6.54
4.75


15.7
13.7
0.02
0.01


5.7
3.6
-
-


  aColumn 1:  Gas-Chrom  Q   (80/100  mesh)  coated  with 1.95% QF-1/1.5% OV-17
packed 1n a 1.2-m x 2-mm or  4-mm  I.D.  glass column.  A 2-mrn I.D. column and
nitrogen gas at 44 mL/m1n flow  rate were used when determining Isophorone and
nitrobenzene by GC/FID.  The  column  temperature was held Isothermal at 85*C.
A 4-mm I.D. column and  10%  methane/90%  argon  carrier gas at 44 mL/m1n flow
rate were used when  determining  the  dlnltrotoluenes  by GC/ECD.  The column
temperature was held Isothermal at 145*C.

  bColumn 2:  Gas-Chrom Q (80/100 mesh) coated with 3% OV-101 packed 1n a 3.0-
m x 2-mm or 4-mm I.D. glass  column.   A 2-mm I.D. column and nitrogen carrier
gas  at  44  mL/m1n  flow  rate  were  used  when  determining  Isophorone and
nitrobenzene by GC/FID.  The column  temperature was held Isothermal at 100*C.
A 4-mm I.D. column and  10%  methane/90%  argon  carrier gas at 44 ml/mln flow
rate were  used  to  determine  the  dlnltrotoluenes  by  GC/ECD.   The column
temperature was held Isothermal at 150*C.
 TABLE 2.   DETERMINATION  OF  PRACTICAL QUANTITATION  LIMITS  (PQL)  FOR VARIOUS
           MATRICES3


     Matrix                                                    Factorb


 Ground water                                                     10
 Low-level  soil  by sonlcation with GPC  cleanup                   670
 High-level soil and  sludges by sonlcatlon                    10,000
 Non-water mlsdble waste                                    100,000


      aSample PQLs are highly  matrix-dependent.     The  PQLs listed  herein  are
      provided for guidance and may not always be achievable.

      Multiply the  Method  Detection   Limits  1n  Table  1  by  the Factor to
      determine the PQL for each analyte In the matrix to be analyzed.
                                   8090 - 2
                                                          Revision
                                                          Date  September 1986

-------
4.0  APPARATUS AND MATERIALS
     4.1  Gas chromatograph;
          4.1.1  Gas  chromatograph:    Analytical   system  complete  with gas
     chromatograph  suitable  for   on-column   Injections  and  all  required
     accessories, including detectors,  column  supplies, recorder, gases, and
     syringes.  A data system for  measuring peak areas and/or peak heights is
     recommended.
          4.1.2  Columns:
               4.1.2.1  Column 1:  1.2-m x 2- or 4-mrn I.D. glass column packed
          with  1.95%  QF-1/1.5%  OV-17  on   Gas-Chrom  Q  (80/100  mesh)  or
          equivalent.
               4.1.2.2  Column 2:  3.0-m x 2- or 4-mm I.D. glass column packed
          with 3% OV-101 on Gas-Chrom Q (80/100 mesh) or equivalent.
          4.1.3  Detectors:  Flame ionization (FID) or electron capture (ECD).
     4.2  Kuderna-Danish (K-D) apparatus;
          4.2.1  Concentrator tube:   10-mL, graduated (Kontes K-570050-1025 or
     equivalent).  Ground-glass  stopper  is  used  to  prevent evaporation of
     extracts
          4.2.2  Evaporation   flask:      500-mL   (Kontes   K-570001-500  or
     equivalent).  Attach to concentrator tube with springs.            *
          4.2.3  Snyder column:    Three-ball  macro  (Kontes K-503000-0121 or
     equivalent).
          4.2.4  Snyder  column:    Two-ball  micro  (Kontes  K-569001-0219 or
     equivalent).
     4.3  Boiling chips;  Solvent extracted, approximately 10/40 mesh  (silicon
 carbide or  equivalent).
     4.4  Water  bath;    Heated,  with  concentric  ring  cover,   capable  of
 temperature control  (+5*C).  The bath should be used in  a hood.
     4.5  Volumetric  flasks;  10-, 50-, and 100-mL, ground-glass stopper.
     4.6  Microsyringe;  10-uL.
     4.7  Syringe;   5-mL.
     4.8  Vials;  Glass, 2-,  10-,  and  20-mL capacity with Teflon-lined  screw
 cap.
                                   8090  -  3
                                                          Revision
                                                          Date   September  1986

-------
5.0  REAGENTS

     5.1  Solvents;  hexane, acetone (pesticide quality or equivalent.)

     5.2  Stock standard solutions;

          5.2.1  Prepare stock standard solutions  at  a concentration of 1.00
     ug/uL by dissolving 0.0100 g of  assayed reference material  1n hexane and
     diluting to volume 1n a  10-mL  volumetric  flask.  Larger volumes can be
     used at the convenience of the  analyst.  When compound purity 1s assayed
     to be 96%  or  greater,  the  weight  can  be  used without correction to
     calculate the concentration of the stock standard.  Commercially prepared
     stock standards can be used at any concentration 1f they are certified by
     the manufacturer or by an Independent source.

          5.2.2  Transfer  the  stock  standard  solutions  Into Teflon-sealed
     screw-cap bottles.  Store at  4*C and protect from light.  Stock standards
     should  be checked  frequently for  signs  of degradation or evaporation,
     especially just prior to preparing calibration standards from them.

          5.2.3  Stock  standard solutions must be  replaced after one year, or
     sooner  1f comparison with check standards Indicates a problem.

     5.3  Calibration standards;   Calibration  standards  at a minimum of  five
 concentration  levels are prepared  through dilution of the stock standards  with
 hexane.   One of the  concentration levels   should be at a concentration near,
 but  above, the method   detection   limit.     The remaining concentration levels
 should  correspond  to   the   expected  range  of  concentrations  found 1n  real
 samples or should  define the working  range  of the GC.  Calibration  solutions
 must be replaced   after six months,  or  sooner  1f  comparison with a check
 standard  indicates a problem.

     5.4   Internal standards (if  internal  standard  calibration is used); To
 use  this  approach, the  analyst must select one or more Internal standards  that
 are  similar  In analytical behavior to  the compounds of  interest.  The analyst
 must further demonstrate that  the measurement  of the  Internal standard is not
 affected  by  method or matrix interferences.    Because  of these limitations, no
 internal  standard  can be suggested that  is applicable  to all samples.

           5.4.1   Prepare  calibration    standards    at   a   minimum  of   five
     concentration levels   for  each  parameter  of  interest  as  described in
     Paragraph 5.3.

           5.4.2   To  each calibration  standard,  add   a  known  constant  amount of
     one  or  more  internal  standards,  and dilute  to volume with hexane.

           5.4.3   Analyze  each calibration  standard according to Section  7.0.

      5.5  Surrogate standards;  The analyst   should  monitor the performance of
 the extraction,  cleanup(when  used),   and   analytical   system and  the  effec-
 tiveness  of the  method  in  dealing  with   each  sample matrix by  spiking each
                                   8090 - 4
                                                          Revision      0
                                                          Date  September 1986

-------
sample,  standard, and reagent water blank with one or two surrogates (e.g.,  2-
fluoroblphenyl) recommended to encompass the  range of the temperature program
used 1n  this method.   Method  3500,  Section 5.3.1.1,  details Instructions  on
the preparation of base/neutral  surrogates.    Deuterated analogs of analytes
should not be  used  as  surrogates  for  gas  chromatographic analysis due  to
coelutlon problems.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     6.1  See the Introductory  material  to  this  chapter, Organic Analytes,
Section 4.1.  Extracts must be  stored under refrigeration and analyzed within
40 days of extraction.


7.0  PROCEDURE

     7.1  Extraction;

          7.1.1  Refer to Chapter Two for guidance on choosing the appropriate
     extraction  procedure.   In general,  water  samples  are extracted at a pH
     between 5 to 9 with methylene  chloride, using either Method 3510 or 3520.
     Solid  samples are extracted using either Method 3540 or 3550.

          7.1.2  Prior to gas  chromatographic analysis,  the extraction solvent
     must be exchanged to hexane.    The  exchange is performed during the  K-D
     procedures  listed in all  of the  extraction methods.  The exchange may be
     performed in one of two ways,  depending on the data requirements.  If  the
     detection limits cited  in Table  1  must be achieved, the exchange should
     be  performed  as described starting  in  Section 7.1.4.  If these  detection
     limits are  not necessary,  solvent  exchange  is performed as outlined in
     Section 7.1.3.

          7.1.3  Solvent exchange when   detection  limits   In  Table  1 are  not
     required:

               7.1.3.1  Following K-D of the methylene chloride extract to
          1 mL using the macro-Snyder column,  allow the apparatus  to cool  and
          drain  for at least 10 min.

               7.1.3.2  Momentarily remove  the  Snyder   column,   add   50 mL of
          hexane,  a new boiling  chip,   and  reattach the macro-Snyder column.
          Concentrate the extract using  1   mL  of  hexane to prewet the Snyder
          column.   Place the  K-D   apparatus  on   the  water  bath  so that  the
          concentrator tube  is partially Immersed  in   the hot water.  Adjust
          the  vertical position of  the  apparatus and the water temperature, as
          required, to complete concentration in 5-10 min.  At the  proper rate
          of distillation the  balls of   the column will actively chatter,  but
          the  chambers will  not  flood.     When  the apparent volume  of liquid
          reaches  1 mL, remove the   K-D  apparatus  and  allow it to  drain  and
          cool for  at  least  10 min.    The  extract will  be  handled  differently
                                   8090  -  5
                                                         Revision      0
                                                         Date  September  1986

-------
    at this point, depending on  whether  or  not cleanup is needed.  If
    cleanup is not required, proceed  to  Paragraph 7.1.3.3.  If cleanup
    is needed, proceed to Paragraph 7.1.3.4.

         7.1.3.3  If cleanup of the extract  is not required, remove the
    Snyder column and  rinse  the  flask  and  its  lower joint into the
    concentrator tube  with  1-2  mL  of  hexane.    A  5-mL  syringe is
    recommended for this operation.  Adjust the extract volume to
    10.0 ml.  Stopper  the  concentrator  tube and store refrigerated at
    4*C if further processing will not be performed immediately.  If the
    extract  will  be  stored  longer   than  two  days,  it  should  be
    transferred to a  Teflon-sealed  screw-cap  vial.   Proceed with gas
    chromatographic analysis.

         7.1.3.4  If cleanup  of  the  extract  is  required, remove the
    Snyder column and  rinse  the  flask  and  its  lower joint into the
    concentrator  tube with  a minimum  amount  of hexane.  A  5-mL syringe
    is recommended for this operation.   Add a clean boiling chip to the
    concentrator  tube and attach a two-ball micro-Snyder column.  Prewet
    the column by adding about 0.5 ml  of   hexane to the top.   Place the
    micro-K-D apparatus on  the  water  bath  (80*C)  so that the concen-
    trator tube  is partially   immersed   in  the  hot  water.  Adjust the
    vertical position of  the  apparatus  and  the water temperature, as
    required, to  complete concentration  in  5-10 min.  At the proper rate
    of distillation  the balls  of  the  column will actively  chatter, but
    the chambers  will not   flood.    When   the apparent volume  of liquid
    reaches  0.5  ml,  remove  the K-D   apparatus  and allow it  to  drain and
    cool  for at  least  10 min.

          7.1.3.5  Remove  the micro-Snyder  column  and  rinse  the  flask and
     its  lower joint  into  the   concentrator  tube   with  0.2  ml  of hexane.
    Adjust the  extract  volume  to  2.0 ml  and proceed  with Method 3620.

     7.1.4  Solvent exchange when  detection limits listed 1n Table 1 must
be achieved:

          7.1.4.1  Following K-D of the methylene chloride  extract to
     1 ml using the macro-Snyder column,   allow the apparatus to cool  and
     drain for at least 10 min.

          7.1.4.2  Remove the Snyder column  and   rinse  the flask and its
     lower joint into  the  concentrator  tube  with  1-2 ml of methylene
     chloride.  A 5-mL syringe  is  recommended  for this operation.  Add
     1-2 mL of hexane,  a clean boiling  chip, and attach  a two-ball  micro-
     Snyder column.  Prewet the column  by  adding 0.5 mL of hexane to the
     top.   Place the micro-K-D  apparatus  on the water bath (60-65*C)  so
     that the concentrator  tube is  partially  immersed  in the hot water.
     Adjust  the  vertical   position  of  the  apparatus  and  the  water
     temperature, as required,  to complete concentration in 5-10 min.  At
     the proper  rate  of   distillation  the  balls  of  the  column will
     actively chatter,   but  the  chambers  will   not  flood.    When the
     apparent volume of liquid reaches  0.5  mL,  remove  the K-D apparatus
     and allow it to drain  and cool  for at least 10 min.

                             8090 - 6
                                                    Revision      0
                                                    Date  September 1986

-------
               7.1.4.3  Remove the mlcro-Snyder column  and rinse the flask and
          its lower joint into the concentrator  tube with a minimum amount of
          hexane.  The volume of the  extract  should be adjusted to 1.0 ml If
          the extract will  be analyzed  without  cleanup.   If the extract will
          require cleanup,  adjust the volume  to  2.0 ml with hexane.   Stopper
          the concentrator  tube  and  store  refrigerated  at  4*C if further
          processing will not be performed  immediately.   If the extract will
          be stored longer  than  two  days,  it  should  be  transferred to a
          Teflon-sealed screw-cap vial.    Proceed  with  either gas chromato-
          graphic analysis or with cleanup,  as necessary.

     7.2  Gas chromatography conditions  (Recommended);   The determination of
dinitrotoluenesshouldbe  performedusingGC/ECD.     All  other compounds
amenable to this method are to be analyzed by GC/FID.

          7.2.1  Column 1:   Set  10%  methane/90%  argon  carrier gas flow at
     44 mL/min flow rate.  For a 2-mm I.D. column, set  the temperature at 85*C
     isothermal.   For  a  4-mm  I.D.  column,  set  the  temperature at 145*C
     isothermal.

          7.2.2  Column 2:   Set  10%  methane/90%  argon  carrier gas flow at
     44 mL/min flow rate.   For  a  2-mm  I.D.  column, set the temperature at
     100*C isothermal.  For a 4-mm  I.D.  column, set the temperature at 150*C
     isothermal.

     7.3  Calibration:    Refer   to   Method   8000  for  proper  calibration
techniques^Use Table 1 and especially  Table 2 for guidance on selecting the
lowest point on the calibration curve.

          7.3.1  The procedure for  internal  or external standard calibration
     may be used.  Refer to  Method  8000   for  a description of each of these
     procedures.

          7.3.4  If cleanup  is performed  on  the  samples,  the analyst should
     process a series  of  standards   through  the   cleanup  procedure and  then
     analyze the samples by  GC.    This  will confirm  elution patterns and the
     absence of  interferents from the  reagents.

     7.4  Gas  chromatographic analysis;

          7.4.1  Refer to Method 8000.    If  the  internal standard calibration
     technique  is  used,  add  10  uL of   internal standard to  the  sample prior to
     injection.

          7.4.2  Follow  Section  7.6   in   Method   8000   for  instructions on the
     analysis  sequence,  appropriate   dilutions,  establishing  daily  retention
     time windows, and identification  criteria.   Include a mid-level standard
     after  each  group of 10  samples   in  the analysis  sequence when  using FID
     and after each group of 5  samples  in  the analysis  sequence when using
     ECD.
                                  8090 - 7
                                                         Revision      0
                                                         Date  September 1986

-------
          7.4.3   An  example   of   a  GC/FID  chromatogram  for nitrobenzene and
     isophorone  is shown  in  Figure  1.    Figure  2   is an example of a GC/ECD
     chromatogram of the  dinitrotoluenes.

          7.4.4   Record the   sample  volume   injected and   the  resulting peak
     sizes (in area  units or peak heights).

          7.4.5   Using either the  internal   or external  calibration procedure
     (Method 8000),  determine the identity  and  quantity of each analyte peak
     in  the  sample  chromatogram.    See  Section   7.8  of Method  8000 for
     calculation equations.

          7.4.6   If  peak detection and  identification   are prevented due  to
     interferences,  the hexane extract may  undergo cleanup using Method 3620.

     7.5  Cleanup;

          7.5.1   Proceed with  Method  3620,   using   the  2-mL  hexane extracts
     obtained from either Paragraph 7.1.3.5 or 7.1.4.3.

          7.5.2   Following cleanup, the extracts  should   be  analyzed by GC,  as
     described in the previous paragraphs and in  Method 8000.


8.0  QUALITY CONTROL

     8.1  Refer  to   Chapter  One  for  specific   quality control  procedures.
Quality control  to validate sample extraction is  covered  in  Method  3500  and  in
the extraction method utilized.   If  extract  cleanup was  performed,  follow the
QC in Method 3600 and in the specific cleanup method.
     8.2  Procedures to check  the  GC
8000, Section 8.6.
                                        system  operation  are found in Method
          8.2.1  The quality control  check  sample  concentrate (Method 8000,
     Section 8.6) should contain each  parameter  of  interest in acetone at a
     concentration of 20  ug/mL  for  each  dinitrotoluene  and  100 ug/mL for
     isophorone and nitrobenzene.

          8.2.2  Table 3 indicates the  calibration and QC acceptance criteria
     for this  method.    Table  4  gives  method  accuracy  and  precision as
     functions of concentration for the analytes of interest.  The contents of
     both Tables should be used to  evaluate a laboratory's ability to perform
     and generate acceptable data by this method.

     8.3  Calculate surrogate  standard  recovery  on  all samples, blanks, and
 spikes.  Determine if  the  recovery  is  within limits  (limits established by
 performing  QC procedures outlined in Method 8000, Section 8.10).
                                   8090  -  8
                                                          Revision      0
                                                          Date  September  1986

-------
             COLUMN: 1.5% OV-17 +1JS% QF-1
                      ON GAS CHROM Q
             TEMPERATURE: 8S°C.
             DETECTOR: FLAME IONIZATION
                 ui
                 5
                     i
                     o
            24   6   8   10  12
            RETENTION TIME-MINUTES
Figure 1. Gas chromatogram of nitrobcnzcna and itophorone.
                     8090 - 9
                                           Revision      p
                                           Date  September 1986

-------
          COLUMN: 1.5% OV-17+1.95% OF-1
                  ON GAS CHROM Q
          TEMPERATURE: 145®C.
          DETECTOR:  ELECTRON CAPTURE
             2

             1
             o
             
-------
          8.3.1   If recovery 1s not within limits,  the  following 1s  required.

               •   Check to  be  sure  there  are  no errors  1n calculations,
                  surrogate solutions  and  Internal standards.  Also,  check
                  Instrument performance.

               •   Recalculate the data and/or reanalyze  the extract 1f any of
                  the above checks reveal  a problem.

               •   Reextract and reanalyze the sample  1f none of the above are
                  a problem or flag the data as "estimated concentration."


9.0  METHOD PERFORMANCE

     9.1  The method  was  tested  by  18  laboratories  using  reagent water,
drinking water, surface water, and  three industrial wastewaters spiked at six
concentrations over the range  1.0  to  515  ug/L.  Single operator precision,
overall precision, and method accuracy  were  found  to be directly related to
the concentration of the parameter  and  essentially independent of the sample
matrix.    Linear  equations  to  describe  these   relationships  for  a flame
ionizatlon detector are presented in Table 4.

     9.2  The accuracy and precision obtained will  be determined by the sample
matrix, sample-preparation technique, and calibration procedures used.


10.0  REFERENCES

1.  "Development and Application of Test  Procedures for Specific Organic Toxic
Substances in  Wastewaters.    Category  4  -   Nitroaromatlcs and Isophorone,'
Report for EPA Contract 68-03-2624  (in preparation).

2.  "Determination  of  Nitroaromatics    and    Isophorone   in   Industrial  and
Municipal  Wastewaters,"    EPA-600/4-82-024,    U.S.  Environmental  Protection
Agency, Environmental  Monitoring   and  Support Laboratory,  Cincinnati, Ohio
45268, June 1982.

3.  Burke, J.A.   "Gas  Chromatography  for  Pesticide  Residue  Analysis; Some
Practical  Aspects,"  Journal  of   the  Association  of  Official  Analytical
Chemists, 48, 1037, 1965.

4.  "EPA  Method   Validation   Study   19,   Method 609  (Nitroaromatlcs  and
Isophorone)," Report for  EPA  Contract 68-03-2624  (1n preparation).

5.  U.S.  EPA 40 CFR Part  136,  "Guidelines Establishing Test  Procedures for the
Analysis  of Pollutants Under  the Clean Water Act;  Final Rule and Interim Final
Rule and  Proposed  Rule,"  October 26,  1984.

6.  Provost, L.P.  and R.S.  Elder,   "Interpretation of Percent  Recovery Data,"
American  Laboratory, lj>,  pp.  58-63,  1983.
                                   8090  -  11
                                                          Revision
                                                          Date   September  1986

-------
TABLE 3.  QC ACCEPTANCE CRITERIA*


Parameter
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Isophorone
Nitrobenzene
Test
cone.
(ug/L)
20
20
100
100
Limit
for s
(ug/L)
5.1
4.8
32.3
33.3
Range
for X
(ug/L)
3.6-22.8
3.8-23.0
8.0-100.0
25.7-100.0
Range
P, PS
(%)
6-125
8-126
D-117
6-118
     s = Standard deviation of four recovery measurements, 1n ug/L.
     7 = Average recovery for four recovery measurements, 1n ug/L.
     P, Ps = Percent recovery measured.
     D = Detected; result must be greater than zero.
     aCr1teria from 40 CFR Part 136 for  Method 609.  These criteria are based
directly upon the method performance  data  1n  Table 4.  Where necessary, the
limits for recovery have been broadened  to assure applicability of the limits
to concentrations below those used to develop Table 4.
                                   8090 - 12
                                                          Revision      0
                                                          Date  September 1986

-------
TABLE 4.  METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION
Parameter
2,4-D1n1trotoluene
2,4-D1n1trotoluene
Isophorene
Nitrobenzene
Accuracy, as
recovery, x1
(ug/L)
0.65C+0.22
0.66C+0.20
0.49C+2.93
0.60C+2.00
Single analyst
precision, sr'
(ug/L)
0.207+0.08
0.197+0.06
0.287+2.77
0.257+2.53
Overal 1
precision,
S' (ug/L)
0.377-0.07
0.367-0.00
0.467+0.31
0.377-0.78
     x1  = Expected  recovery  for  one  or  more  measurements  of  a  sample
           containing a concentration of C, 1n ug/L.

     sr' = Expected single analyst  standard  deviation  of measurements at an
           average concentration of 7, 1n ug/L.

     S1  = Expected Interlaboratory standard  deviation  of measurements at an
           average concentration found of 7, 1n ug/L.

     C   = True value for the concentration, 1n ug/L.

     7   = Average recovery found for measurements of samples containing a
           concentration of C, In ug/L.
                                  8090 - 13
                                                         Revision
                                                         Date  September 1986

-------
                                          METHOD B090

                               NITHOAHOMATICS AND CYCLIC KETONES
                           7.1.1
                                  Choose
                                extraction
                            procedure  from
                                Chapter 2
                                                    7.1.4
                           Concentrate to
                           1 mL using K-D
                             apparatus
 7.1.3
	1  Rinse
    with hexane:
  re-concentrate
     to  .5  ml_:
  adjust to Z  mL
 7.1.3
Is cleanup
 required?
  Cleanup using
   Method 362O



7.1.31 Rinse
	 1 flesh
Into concentra-
tor tube with
hexane: adjust
to 10 mL
                              0
                                                           Rinse
                                                     .  with hexane:
                                                      concetrate to
                                                    .5 ml using K-D
                                                                    Yes

7.1.4



Adjust volume
to
1 mL
                                                                              7.1.4
Adjust volume
   to Z mL
                                                                              7.1.4
                                                  Cleanup using
                                                   Method 362O
                                      8090 -  14
                                                                Revision       Q
                                                                Date  September  1986

-------
          METHOD BO90

NITHOAROMATICS AND CYCLIC KETONES
           (Continued)
         Set GC column
           Operating
           conditions
7.3

Calibrate (see
Method 8000)


7.4 j
Perform
GC analysis
(see Method
aoooi
       (     Stop      J
    8090  -  15
                              Revision       p
                              Date  September 1986

-------
                                 METHOD 8100

                      POLYNUCLEAR AROMATIC HYDROCARBONS
1.0  SCOPE AND APPLICATION

     1.1  Method 8100  is  used  to  determine  the  concentration  of certain
polynuclear aromatic hydrocarbons (PAH).  Table 1 indicates compounds that may
be determined by this method.

     1.2  The packed column gas  chromatographic  method described here cannot
adequately resolve the  following  four  pairs  of  compounds:  anthracene and
phenanthrene;  chrysene   and   benzo(a)anthracene;  benzo(b)fluoranthene  and
benzo(k)fluoranthene; and  dibenzo(a,h)anthracene  and indeno(l,2,3-cd)pyrene.
The use of a capillary column instead  of the packed column, also described in
this method, may adequately resolve  these  PAHs.  However, unless the purpose
of the  analysis  can  be  served  by  reporting  a  quantitative  sum  for an
unresolved PAH pair, either liquid chromatography  (Method 8310) or gas chroma-
tography/mass spectroscopy (Method 8270) should be used for these compounds.


2.0  SUMMARY OF METHOD

     2.1  Method  8100  provides   gas   chromatographic  conditions  for  the
detection of ppb levels of  certain  polynuclear aromatic hydrocarbons.  Prior
to use  of this method, appropriate   sample extraction techniques must be used.
Both neat and diluted  organic  liquids  (Method   3580, Waste Dilution) may be
analyzed by direct  injection.  A 2-  to  5-uL aliquot of the  extract 1s Injected
into a  gas chromatograph  (GC) using  the  solvent flush technique, and compounds
in the  GC effluent  are detected by a flame ionization detector  (FID).

     2.2  If   interferences   prevent proper   detection  of  the  analytes  of
interest, the method may   also  be   performed  on  extracts  that have undergone
cleanup using  silica gel  column cleanup  (Method 3630).


3.0  INTERFERENCES

     3.1  Refer to  Methods 3500, 3600,  and 8000.

     3.2  Solvents,  reagents, glassware, and   other sample  processing hardware
may   yield   discrete     artifacts     and/or   elevated    baselines   causing
misinterpretation of gas chromatograms.    All   of  these  materials must be
demonstrated to  be free from  interferences,  under  the  conditions of the
analysis, by analyzing  method  blanks.    Specific  selection  of reagents and
purification of solvents  by distillation in all-glass systems may be required.

     3.3  Interferences coextracted  from samples  will vary  considerably from
source  to source, depending upon  the   waste   being sampled.  Although general
cleanup techniques  are recommended as part  of this method, unique samples may
require additional  cleanup.


                                  8100 - 1
                                                         Revision      0
                                                         Date   September 1986

-------
TABLE 1.  GAS CHROMATOGRAPHY OF POLYNUCLEAR AROMATIC HYDROCARBONS3

Compound                                    Retention time (min)
Acenaphthene                                        10.8
Acenaphthylene                                      10.4
Anthracene                                          15.9
Benzo(a)anthracene                                  20.6
Benzo(a)pyrene                                      29.4
Benzo
Benzo
Benzo
Benzo
      b)fluoranthene                                28.0
      j)fluoranthene
      kjfluoranthene                                28.0
      ghi)perylene                                  38.6
Chrysene                                            24.7
Dibenz(a,h)acridine
Dibenz(a,j)acridine
Dibenzo(a,h)anthracene                              36.2
7H-Dibenzo(c,g)carbazole
Dibenzo(a,e)pyrene
Dibenzo(a,h)pyrene
Di benzo(a, i)pyrene
Fluoranthene                                        19.8
Fluorene                                            12.6
Indeno(l,2,3-cd)pyrene                              36.2
3-Methylcholanthrene
Naphthalene                                          4.5
Phenanthrene                                        15.9
Pyrene                                              20.6
      aResu1ts  obtained using Column  1.
                                   8100 - 2
                                                          Revision
                                                          Date  September 1986

-------
4.0  APPARATUS AND MATERIALS

     4.1  Gas chrornatograph;

          4.1.1  Gas  chromatograph:    Analytical   system  complete  with gas
     chromatograph  suitable  for   on-column   injections  and  all  required
     accessories, including detectors,  column  supplies, recorder, gases, and
     syringes.  A data system for  measuring peak areas and/or peak heights is
     recommended.

          4.1.2  Columns:

               4.1.2.1  Column 1:  1.8-m x  2-mm I.D. glass column packed with
          3% OV-17 on Chromosorb W-AW-DCMS (100/120 mesh) or equivalent.

               4.1.2.2  Column 2:   30-m  x  0.25-mm  I.D.  SE-54 fused silica
          capillary column.

               4.1.2.3  Column 3:   30-m  x  0.32-mm  I.D.  SE-54 fused silica
          capillary column.

          4.1.3  Detector:  Flame ionizatlon (FID).

     4.2  Volumetric flask;  10-, 50-, and 100-mL,  ground-glass stopper.

     4.3  Microsyrlnge:  10-uL.


5.0  REAGENTS

     5.1  Solvents:    Hexane,   isooctane  (2,2,4-trimethylpentane) (pesticide
quality or equivalent).

     5.2  Stock  standard solutions;

          5.2.1  Prepare stock standard solutions  at  a  concentration of  1.00
     ug/uL by  dissolving 0.0100  g  of assayed reference  material  1n isooctane
     and diluting to volume 1n a 10-mL  volumetric  flask.  Larger volumes can
     be used at  the  convenience of  the  analyst.    When compound purity is
     assayed to  be 96% or  greater,  the  weight can  be used without correction
     to calculate  the   concentration of  the  stock  standard.   Commercially
     prepared  stock standards can  be used  at  any concentration 1f they are
     certified by the manufacturer or by an  Independent  source.

          5.2.2  Transfer   the   stock standard  solutions  Into  Teflon-sealed
     screw-cap bottles.  Store at 4*C and  protect  from light.  Stock standards
     should  be checked   frequently  for  signs  of degradation or evaporation,
     especially  just prior to preparing calibration  standards  from them.

          5.2.3  Stock standard  solutions  must be  replaced after one year, or
     sooner  if comparison  with check  standards Indicates  a problem.
                                   8100 -  3
                                                          Revision
                                                         Date  September 1986

-------
     5.3  Calibration standards;  Calibration  standards  at a minimum of five
concentrationlevelsshouldbe  prepared  through  dilution  of  the  stock
standards with isooctane.   One  of  the  concentration  levels should be at a
concentration near, but  above,  the  method  detection  limit.  The remaining
concentration levels should correspond to the expected range of concentrations
found in real samples or should  define  the  working  range of the GC.  Cali-
bration solutions must be replaced  after  six months, or sooner if comparison
with a check standard indicates a problem.

     5.4  Internal standards (if internal  standard  calibration is used):  To
use this approach, the analyst must select one or more internal standards that
are similar  in analytical behavior to  the compounds of interest.  The analyst
must further demonstrate that the measurement  of the internal standard is not
affected by  method or matrix interferences.   Because of these limitations, no
internal standard can be suggested that  is applicable to all samples.

          5.4.1  Prepare  calibration    standards   at   a   minimum  of  five
     concentration   levels  for  each  analyte  of  interest  as  described in
     Paragraph 5.3.

          5.4.2  To  each calibration standard, add  a known constant amount of
     one or  more internal standards, and dilute to volume with isooctane.

          5.4.3  Analyze each calibration standard according to Section 7.0.

     5.5  Surrogate  standards:  The analyst  should monitor the performance of
the  extraction,cleanup(when   used),    and   analytical  system  and  the
effectiveness of the method in dealing with  each sample matrix by spiking each
sample, standard, and reagent water blank with one or two surrogates  (e.g., 2-
fluorobiphenyl and 1-fluoronaphthalene)  recommended  to encompass the range of
the temperature program  used in  this  method.   Method 3500, Section 5.3.1.1,
details  instructions    on   the   preparation   of  base/neutral  surrogates.
Deuterated analogs of  analytes  should  not be  used  as  surrogates for gas
chromatographic analysis due to coelution problems.


6.0  SAMPLE  COLLECTION,  PRESERVATION, AND HANDLING

     6.1  See the  introductory  material  to this  chapter,  Organic  Analytes,
Section  4.1. Extracts must be  stored  under  refrigeration and  must be  analyzed
within  40 days of  extraction.


7.0   PROCEDURE

      7.1   Extraction;

           7.1.1   Refer to  Chapter  Two  for guidance  on choosing the  appropriate
      extraction  procedure.    In  general,   water  samples   are  extracted at a
      neutral pH  with methylene  chloride,   using  either Method 3510 or 3520.
      Solid  samples are extracted using either Method  3540 or 3550.   To achieve
      maximum sensitivity with  this method,  the  extract must be concentrated to
      1  mL.

                                   8100 - 4
                                                          Revision      0
                                                          Date  September 1986

-------
     7.2   Gas  chromatography  conditions  (Recommended);

          7.2.1   Column 1:  Set  nitrogen   carrier  gas  flow at 40-mL/m1n flow
     rate.  Set  column temperature  at  100*C for 4 m1n; then program  at 8*C/min
     to a final  hold at 280*C.

          7.2.2   Column 2:  Set  helium  carrier  gas at  20-cm/sec  flow rate.
     Set  column  temperature at  35'C  for 2  min;   then  program at 10*C/m1n  to
     265*C and hold for 12  m1n.

          7.2.3   Column 3:  Set  helium  carrier  gas at  60 cm/sec  flow rate.
     Set  column  temperature at  35*C  for 2  m1n;   then  program at 10*C/m1n  to
     265*C and hold for 3 mln.

     7.3   Calibration:    Refer   to  Method  8000   for   proper calibration
techniques.

          7.3.1  The procedure for  Internal  or  external  standard calibration
     may be used.  Refer to  Method  8000   for  a description  of  each of these
     procedures.

          7.3.2  If cleanup is performed on  the  samples,  the analyst  should
     process a series  of  standards  through  the   cleanup procedure and  then
     analyze the samples by GC.   This  will validate elutlon  patterns and  the
     absence of interferents  from the reagents.

     7.4  Gas chromatographic analysis;

          7.4.1  Refer to Method 8000.    If the  internal  standard calibration
     technique is used, add 10 uL of  internal standard  to the sample prior to
     injection.

          7.4.2  Follow Section 7.6  in   Method  8000  for instructions  on  the
     analysis sequence,  appropriate  dilutions,   establishing daily retention
     time windows, and Identification criteria.    Include  a mid-level standard
     after each group of 10 samples in the  analysis sequence.

          7.4.3  Record the  sample  volume  Injected  and  the  resulting  peak
     sizes (in area units or peak heights).

          7.4.4  Using either the  internal  or external calibration procedure
     (Method 8000), determine the identity  and quantity of each component peak
     in the sample chromatogram  which  corresponds  to the compounds used for
     calibration purposes.  See  Section  7.8  of  Method  8000 for calculation
     equations.

          7.4.5  If peak detection  and  identification  are  prevented due to
     interferences, the extract may undergo cleanup using  Method 3630.

     7.5  Cleanup:

          7.5.1  Proceed with Method  3630.    Instructions  are given in this
     method for exchanging the solvent of the extract to hexane.

                                  8100 - 5
                                                         Revision      0
                                                         Date  September 1986

-------
          7.5.2  Following cleanup,  the extracts should  be analyzed by GC,  as
     described in the previous paragraphs and in Method 8000.


8.0  QUALITY CONTROL

     8.1  Refer  to  Chapter  One  for  specific  quality  control procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If  extract cleanup was performed, follow the
QC in Method 3600 and in the specific cleanup method.

     8.2  Procedures to check  the  GC  system  operation  are found in Method
8000, Section 8.6.

          8.2.1  The quality control  check  sample  concentrate  (Method 8000,
     Section 8.6) should contain each  analyte at the following concentrations
     in acetonitrile:    naphthalene,  100  ug/mL;  acenaphthylene, 100 ug/mL;
     acenaphthene, 100 ug/mL;  fluorene,  100  ug/mL; phenanthrene, 100 ug/mL;
     anthracene, 100 ug/mL; benzo(k)fluoranthene,  5  ug/mL; and  any other  PAH
     at 10  ug/mL.

          8.2.2  Table 2  indicates  the   calibration and QC acceptance  criteria
     for  this  method.     Table  3  gives  method  accuracy  and   precision as
     functions of concentration  for the  analytes of interest.  The contents of
     both Tables should be used  to  evaluate a  laboratory's ability to perform
     and  generate acceptable  data by  this method.

     8.3  Calculate  surrogate standard   recovery  on   all  samples,  blanks,  and
 spikes.   Determine  if  the recovery   is within limits  (limits established by
 performing  QC  procedures  outlined in  Method  8000, Section  8.10).

          8.3.1   If  recovery  is  not   within  limits,  the  following procedures
     are  required.

                •  Check to be  sure   there  are  no   errors   in  calculations,
                  surrogate solutions  and   internal   standards.    Also,  check
                  instrument  performance.

                •  Recalculate the data and/or reanalyze  the  extract  if any of
                  the  above checks  reveal a  problem.

                •  Reextract and  reanalyze the sample   if none  of  the  above  are
                  a  problem or flag the data as "estimated concentration."


 9.0  METHOD PERFORMANCE

     9.1   The method  was  tested   by  16   laboratories  using  reagent water,
 drinking  water,  surface  water, and   three  industrial  wastewaters  spiked at six
 concentrations over the  range  0.1   to  425  ug/L.   Single operator precision,
 overall  precision,  and method accuracy  were  found  to  be directly related to
                                   8100 - 6
                                                          Revision      0
                                                          Date  September 1986

-------
the concentration of the  analyte  and  essentially  independent of the sample
matrix.    Linear  equations  to  describe  these  relationships  for  a flame
ionization detector are presented in Table 3.

     9.2  This method has been  tested  for  linearity  of spike recovery from
reagent  water  and  has   been   demonstrated   to  be  applicable  over  the
concentration range from 8 x MDL  to  800  x MDL with the following exception:
benzo(ghi)perylene recovery at 80  x  and  800  x  MDL  were low (35% and 45%,
respectively).

     9.3  The accuracy and precision obtained will be determined by the sample
matrix, sample-preparation technique, and calibration procedures used.

10.0  REFERENCES

1.  "Development and Application of Test Procedures for Specific Organic Toxic
Substances in Wastewaters.  Category 9 - PAHs," Report for EPA Contract 68-03-
2624 (in preparation).

2.  Sauter, A.D., L.D. Betowski, T.R. Smith, V.A. Strickler, R.G. Beimer,
B.N. Colby, and J.E. Wilkinson,  "Fused  Silica Capillary Column GC/MS for the
Analysis of Priority Pollutants," Journal of HRC&CC 4, 366-384, 1981.

3.  "Determination of  Polynuclear  Aromatic  Hydrocarbons  in  Industrial and
Municipal  Wastewaters,"   EPA-600/4-82-025,   U.S.   Environmental   Protection
Agency, Environmental  Monitoring  and  Support   Laboratory,  Cincinnati, Ohio
45268,  September 1982.

4.  Burke, J.A.  "Gas  Chromatography  for   Pesticide  Residue  Analysis; Some
Practical  Aspects,"  Journal  of   the  Association  of  Official   Analytical
Chemists, 48,  1037,  1965.

5.  "EPA  Method  Validation   Study   20,   Method 610   (Polynuclear Aromatic
Hydrocarbons)," Report for EPA Contract 68-03-2624 (in preparation).

6.  U.S.  EPA 40 CFR  Part  136,  "Guidelines Establishing Test Procedures for the
Analysis  of  Pollutants Under  the Clean Water Act;  Final  Rule and  Interim  Final
Rule and  Proposed Rule,"  October 26,  1984.

7.  Provost,  L.P. and R.S.  Elder,   "Interpretation of Percent  Recovery Data,"
American  Laboratory, 15,  pp.  58-63,  1983.
                                   8100 -  7
                                                          Revision      0	
                                                          Date  September 1986

-------
TABLE 2.  QC ACCEPTANCE CRITERIA3
Parameter
Acenaphthene
Acenaphthylene
Anthracene
Benzo (a) anthracene
Benzo(a)pyrene
Benzo (b) f 1 uoranthene
Benzo(ghi)perylene
Benzo (k) f 1 uoranthene
Chrysene
Di benzo (a , h) anthracene
Fl uoranthene
Fluorene
Indeno (1,2, 3-cd) py rene
Naphthalene
Phenanthrene
Pyrene
Test
cone.
(ug/L)
100
100
100
10
10
10
10
5
10
10
10
100
10
100
100
10
Limit
for s
(ug/L)
40.3
45,1
28.7
4.0
4.0
3.1
2.3
2.5
4.2
2.0
3.0
43.0
3.0
40.7
37.7
3.4
Range
for 7
(ug/L)
D-105.7
22.1-112.1
11.2-112.3
3.1-11.6
0.2-11.0
1.8-13.8
D-10.7
D-7.0
D-17.5
0.3-10.0
2.7-11.1
D-119
1.2-10.0
21.5-100.0
8.4-133.7
1.4-12.1
Range
P, PS
(%)
D-124
D-139
D-126
12-135
D-128
6-150
D-116
D-159
D-199
D-110
14-123
D-142
D-116
D-122
D-155
D-140
      s  = Standard deviation of four recovery measurements, in ug/L.

      7  = Average recovery  for four recovery measurements, in ug/L.

      P, Ps  =  Percent  recovery measured.

      D  = Detected;  result  must be greater than zero.

      Criteria from 40  CFR Part  136 for  Method 610.  These criteria are based
 directly upon the method performance   data  in  Table 3.  Where necessary, the
 limits  for  recovery have been broadened  to assure  applicability of the limits
 to concentrations below those used to  develop Table 3.
                                   8100 - 8
                                                          Revision      0
                                                          Date  September 1986

-------
TABLE 3.  METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION
Parameter
Acenaphthene
Acenaphthylene
Anthracene
Benzo (a) anthracene
Benzo (a)pyrene
Benzo (b) f 1 uoranthene
Benzo (ghi)perylene
Benzo (k) f 1 uoranthene
Chrysene
Di benzo (a , h) anthracene
Fl uoranthene
Fluorene
Ideno (1 , 2 , 3-cd) pyrene
Naphthalene
Phenanthrene
Pyrene
Accuracy, as
recovery, x1
(ug/L)
0.52C+0.54
0.69C-1.89
0.63C-1.26
0.73C+0.05
0.56C+0.01
0.78C+0.01
0.44C+0.30
0.59C+0.00
0.77C-0.18
0.41C-0.11
0.68C+0.07
0.56C-0.52
0.54C+0.06
0.57C-0.70
0.72C-0.95
0.69C-0.12
Single analyst
precision, sr'
(ug/L)
0.397+0.76
0.367+0.29
0.237+1.16
0.287+0.04
0.387-0.01
0.217+0.01
0.257+0.04
0.447-0.00
0.327-0.18
0.247+0.02
0.227+0.06
0.447-1.12
0.297+0.02
0.397-0.18
0.297+0.05
0.257+0.14
Overall
precision,
S1 (ug/L)
0.537+1.32
0.427+0.52
0.417+0.45
0.347+0.02
0.537-0.01
0.387-0.00
0.587+0.10
0.697+0.10
0.667-0.22
0.457+0.03
0.327+0.03
0.637-0.65
0.427+0.01
0.417+0.74
0.477-0.25
0.427-0.00
     x1  =  Expected  recovery  for  one  or  more  measurements  of  a  sample
            containing a concentration of C, in ug/L.

     sr' =  Expected single analyst  standard  deviation  of measurements at an
            average concentration of 7, in ug/L.

     S1  =  Expected interlaboratory standard  deviation  of measurements at an
            average concentration found of 7, in ug/L.

     C   =  True value for the concentration, in ug/L.

     7   =  Average recovery  found for measurements of samples containing a
            concentration of  C, in ug/L.
                                   8100 -  9
                                                          Revision      0
                                                          Date  September  1986

-------
                                          METHOD 8100

                               POLYNUCLEAR AROMATIC HYDROCARBONS
C
 7.1.1
        Choose
      « appro-
 priate extract-
  Ion procedure
     (refer to
    Chapter S)
  7.2
     Set gas
  chrontatography
   conditions
  7.3
                         through cleanup
                             procedures
                                                     7.4
.Do  GC analysis
   (refer  to
  Method 8000)
        Refer to
     Method 6000
     for proper
     calibration
     techniques
     Q
                                                                              7.5. 1
                            Do cleanup
                           using Method
                              3630
                                      8100  - 10
                                                                 Revision        0	
                                                                 Date  September 1986

-------
                                 METHOD 8110

                       HALOETHERS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   This method covers  the  determination of certain  haloethers.   The
following compounds can be determined by this method:
                                                   Appropriate Technique
Compound Name                  CAS  No.a    3510    3520    3540    3550     3580
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
4-Bromophenyl phenyl ether
4-Chlorophenyl phenyl ether
111-91-1
111-44-4
108-60-1
101-55-3
7005-72-3
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
   a  Chemical  Abstract Services Registry  Number.
   X  Greater than 70 percent recovery by  this technique.

      1.2   This  is  a gas  chromatographic  (GC)   method  applicable  to  the
determination  of  the compounds  listed  above  in  municipal  and  industrial
discharges.  When this method is used to analyze unfamiliar samples for any or
all of the compounds above, compound identifications should be supported by at
least one  additional  qualitative technique.   This method  describes analytical
conditions of a second GC column that can  be used to confirm measurements made
with  the  primary  column.    Method  8270  provides  gas  chromatograph/mass
spectrometer (GC/MS) conditions  appropriate for the qualitative and quantitative
confirmation of results for all of the parameters listed above, using the extract
from this method.

      1.3   The method detection limit  (MDL,  defined  in Section  9.1)  for each
parameter  is listed  in Table  1.  The MDL  for  a specific wastewater may differ
from  that  listed,  depending  upon  the  nature  of  interferences in  the  sample
matrix.

      1.4   This method  is  restricted  to use  by  or under the  supervision of
analysts experienced in the use  of gas chromatography and in the interpretation
of gas  chromatograms.   Each analyst must demonstrate  the  ability  to  generate
acceptable results with this method  using the procedure described in Section 8.2.

      1.5   The toxicity or carcinogenicity  of each  reagent used in this method
has not  been precisely defined.   However, each  chemical  compound should be
treated as a potential health hazard.   From this  viewpoint,  exposure  to these
chemicals  must  be  reduced  to  the lowest  possible level  by whatever  means
available.  The  laboratory  is responsible for maintaining  a  current awareness
file of OSHA regulations  regarding  the safe  handling of the chemicals specified
in this method.  A reference file of material  data handling sheets should also

                                   8110 -  1                          Revision 0
                                                                     July 1992

-------
be made available to all  personnel  involved in the chemical analysis.  Additional
references to laboratory safety are available and have been identified.


2.0   SUMMARY OF METHOD

      2.1   A measured  volume  of  sample,  approximately one-liter,  is  solvent
extracted with  methylene  chloride using a  separatory funnel.   The methylene
chloride extract  is  dried and exchanged to hexane during concentration  to  a
volume of 10 ml or less.  GC conditions are described which permit the separation
and measurement of the compounds in the extract using a  halide specific detector.

      2.2   Method 8110  provides gas chromatographic conditions for the detection
of ppb concentrations of haloethers.   Prior to use of this method,  appropriate
sample extraction techniques must be used.  Both neat and diluted organic liquids
(Method 3580, Waste  Dilution)  may  be  analyzed by direct injection.  A 2 to 5 nL
aliquot  of  the extract is  injected  into a  gas  chromatograph (GC)  using the
solvent  flush  technique,  and  compounds  in  the GC effluent are detected  by an
electrolytic conductivity detector (HECD).


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000.

      3.2   Matrix  interferences   may   be   caused  by  contaminants  that  are
coextracted  from  the sample.   The extent  of matrix  interferences  will  vary
considerably from source to source, depending upon the nature and diversity of
the industrial  complex or municipality being sampled.   The cleanup procedures in
Section  7.3  can  be  used to overcome many  of these interferences,  but unique
samples may require additional cleanup approaches to  achieve the MDL listed in
Table 1.

      3.3   Dichlorobenzenes are known to coelute with haloethers  under  some gas
chromatographic conditions.   If these materials  are present in a sample, it may
be  necessary to analyze  the  extract  with  two  different  column  packings to
completely resolve all  of the compounds.

      3.4   Solvents, reagents, glassware,  and other sample processing hardware
may yield discrete artifacts and/or elevated baselines  causing misinterpretation
of gas chromatograms.   All these materials must be demonstrated to  be free from
interferences under the  conditions of the analysis, by  analyzing reagent blanks.
Specific selection of reagents and purification of solvents by distillation in
all-glass systems may be required.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas   chromatograph  -  An  analytical  system  complete  with
      temperature  programmable   gas   chromatograph   suitable  for  on-column
      injection  and  all  required accessories including  syringes, analytical
      columns,  gases,  detector,  and strip-chart  recorder.   A data system is

                                   8110 - 2                         Revision 0
                                                                     July  1992

-------
      recommended for measuring peak areas.

            4.1.2 Columns

                  4.1.2.1     Column 1 -  1.8  m x 2 mm  ID  pyrex  glass,  packed
            with  Supelcoport,  (100/120  mesh)  coated  with  3%  SP-1000  or
            equivalent.  This column was used  to develop the method performance
            statements  in  Section  9.0.    Guidelines  for the use  of alternate
            column packings are provided in Section 7.3.1.

                  4.1.2.2     Column 2 -  1.8  m x 2 mm  ID  pyrex  glass,  packed
            with  2,6-diphenylene  oxide  polymer  (Tenax-GC  60/80  mesh)  or
            equivalent.

            4.1.3 Detector  -  Electrolytic  conductivity or  microcoulometric.
      These detectors have proven effective in the analysis of wastewaters for
      the parameters listed in the  scope of this method.  The Hall conductivity
      detector (HECD) was used to develop the method performance statements in
      Section 9.0.  Guidelines for the use of alternate detectors are provided
      in Section 7.3.1.  Although less selective, an electron capture detector
      (ECD) is an acceptable alternative.

      4.2   Kuderna-Danish (K-D) apparatus

            4.2.1 Concentrator tube - 10 mL graduated (Kontes K-570050-1025 or
      equivalent).  A  ground  glass stopper is used to  prevent  evaporation of
      extracts.

            4.2.2 Evaporation  flask  -   500  ml   (Kontes   K-570001-0500  or
      equivalent).   Attach  to  concentrator  tube  with springs,  clamps,  or
      equivalent.

            4.2.3 Snyder column  -  Three  ball  macro (Kontes K-503000-0121  or
      equivalent).

            4.2.4 Springs -  1/2 inch (Kontes K-662750 or equivalent).

      4.3   Vials - Amber glass,  10 to 15 ml capacity, with Teflon lined screw-
cap or crimp top.

      4.4   Boiling  chips  -  Approximately  10/40  mesh.   Heat  to  400°C  for
30 minutes or Soxhlet extract with methylene chloride.

      4.5   Water  bath  -  Heated,  with  concentric ring  cover,  capable  of
temperature control (± 2°C).   The bath should  be  used  in a  hood.

      4.6   Balance - Analytical, 0.0001 g.

      4.7   Volumetric flasks, Class A  -  Appropriate sizes  with  ground glass
stoppers.
                                   8110 - 3                         Revision 0
                                                                     July 1992

-------
5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be used in all  tests.  Unless
otherwise indicated, it  is intended that all  inorganic reagents shall conform to
the specifications  of the  Committee  on Analytical  Reagents of  the  American
Chemical Society, where  such specifications  are available.  Other grades may be
used,  provided it is first ascertained  that  the reagent is of sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water  -  All  references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Acetone, CH3COCH3  -  Pesticide quality or equivalent.

      5.4   Hexane, C6HU - Pesticide quality or equivalent.

      5.5   Isooctane, (CH3)3CCH2CH(CH3)2  - Pesticide  quality  or equivalent.

      5.6   Stock standard solutions (1000 mg/L) - Stock standard solutions can
be prepared from pure standard materials or purchased as certified solutions.

            5.6.1 Prepare  stock  standard   solutions  by  accurately  weighing
      0.1000 ± 0.0010 g  of  pure  material.   Dissolve  the  material  in pesticide
      quality acetone and dilute to volume in a 100  ml volumetric flask.  Larger
      volumes can be used at the convenience of the  analyst.   If compound purity
      is certified at 96% or greater, the weight can be used without correction
      to  calculate the  concentration  of  the stock  standard.    Commercially
      prepared stock  standards  can be used  at  any concentration  if  they are
      certified by the manufacturer or by an independent source.

            5.6.2 Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps  or crimp tops.    Store at 4°C and  protect from light.
      Stock  standard solutions  should  be   checked  frequently  for  signs  of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.6.3 Stock standard solutions must be replaced after  six months, or
      sooner if comparison with check standards indicates a problem.

      5.7   Calibration  standards - Calibration standards at a minimum of five
concentrations should be prepared through dilution of the stock standards with
isooctane.  One  of the  concentrations  should be at  a  concentration near, but
above,  the  method  detection  limit.    The  remaining  concentrations  should
correspond  to  the expected range  of concentrations  found in real  samples or
should  define  the  working  range of the GC.   Calibration  solutions  must be
replaced after six months, or sooner if comparison with check standards  indicates
a problem.

      5.8   Internal standards  (if internal  standard calibration  is used) - To
use this approach,  the analyst must select one or more internal standards that
are similar in  analytical behavior to  the  compounds  of interest.  The analyst
must further demonstrate that the  measurement of the internal standard is not
affected by method  or matrix interferences.   Because of these limitations, no
internal standard can be suggested that  is applicable to all  samples.

                                   8110  - 4                         Revision 0
                                                                      July 1992

-------
            5.8.1 Prepare  calibration   standards  at   a   minimum   of   five
      concentrations for each analyte of interest as described in Section 5.7.

            5.8.2 To each calibration standard,  add a known constant amount of
      one or more internal standards, and dilute to volume  with isooctane.

            5.8.3 Analyze each calibration standard according to Section 7.0.

      5.9   Surrogate standards -  The analyst should monitor the performance of
the extraction,  cleanup  (when used), and analytical  system and the effectiveness
of  the  method  in  dealing with  each  sample  matrix by  spiking each  sample,
standard, and reagent blank with one or two surrogates (e.g. haloethers that are
not expected  to  be  in the sample)  recommended  to encompass the range  of the
temperature program  used  in this method.  Method 3500 details instructions on the
preparation of base/neutral  surrogates.  Deuterated analogs of analytes should
not be  used  as  surrogates for  gas chromatographic analysis due  to coelution
problems.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this  chapter,  Organic Analytes,
Section 4.1.   Extracts  must be stored  at  4°C  and analyzed within  40 days of
extraction.
7.0   PROCEDURE

      7.1   Extraction

            7.1.1 Refer to Chapter Two for guidance on choosing the appropriate
      extraction  procedure.    In  general,  water  samples  are  extracted  at  a
      neutral, or as is, pH with methylene chloride,  using either Method 3510 or
      3520.  Solid samples are extracted using either Method 3540 or 3550.

            NOTE: Some of the  haloethers  are very  volatile  and  significant
                  losses  will  occur  in concentration  steps  if  care  is  not
                  exercised.   It  is  important to  maintain  a  constant  gentle
                  evaporation rate and  not to allow  the  liquid volume to fall
                  below 1  to 2 mL before removing  the K-D  apparatus from the hot
                  water bath.

            7.1.2 Prior to gas  chromatographic analysis,  the  extraction solvent
      must be exchanged to  hexane.   The exchange is performed during the K-D
      procedures  listed  in all of  the extraction methods.    The exchange  is
      performed as follows.

                  7.1.2.1      Following K-D of the methylene chloride extract to
            1 mL using the macro-Snyder column, allow the apparatus to cool and
            drain for at least 10 minutes.

                  7.1.2.2     Momentarily remove the  Snyder column,  add 50 mL of
            hexane, a new boiling chip,  and  reattach the macro-Snyder column.
            Concentrate the extract using 1  mL of hexane to prewet  the  Snyder

                                   8110 - 5                         Revision 0
                                                                     July 1992

-------
      column.   Place  the K-D apparatus  on  the water  bath  so that  the
      concentrator tube is partially immersed in the hot water.   Adjust
      the vertical position of the apparatus and the water temperature, as
      required, to complete concentration in 5-10 minutes.  At the proper
      rate of  distillation the balls of the column will  actively chatter,
      but the  chambers will not flood.  When the apparent volume of liquid
      reaches  1  ml, remove the  K-D  apparatus  and allow it to  drain  and
      cool  for  at  least  10 minutes.    The  extract  will  be  handled
      differently  at this point, depending on whether  or not  cleanup is
      needed.  If cleanup is not required, proceed to Section 7.1.2.3.   If
      cleanup  is  needed,  proceed to  Section 7.1.2.4.

            7.1.2.3      If cleanup of the extract is not required, remove
      the Snyder  column and  rinse the  flask and  its lower joint into  the
      concentrator  tube  with  1-2  ml of hexane.   A  5 ml  syringe  is
      recommended  for  this   operation.   Adjust  the extract  volume  to
      10.0 mL.  Stopper the concentrator tube and  store  refrigerated at
      4°C if further processing will  not be performed immediately.  If the
      extract  will   be  stored   longer  than  two  days,  it  should   be
      transferred  to  a  Teflon  lined screw-cap vial.   Proceed with  gas
      chromatographic analysis.

            7.1.2.4     If cleanup of the extract is required, remove  the
      Snyder  column  and  rinse  the  flask and its  lower joint into  the
      concentrator tube with a minimum amount  of  hexane.  A 5 ml  syringe
      is recommended for  this operation.  Add a clean boiling chip to  the
      concentrator tube and attach a two ball micro-Snyder column.  Prewet
      the column  by adding about 0.5 ml  of hexane to the top.   Place  the
      micro-K-D   apparatus  on   the  water  bath   (80°C)   so   that   the
      concentrator tube is partially immersed in the hot water.   Adjust
      the vertical position of the apparatus and the water temperature, as
      required, to complete  concentration in 5-10 minutes.  At the proper
      rate of  distillation the  balls of the column will  actively chatter,
      but the  chambers will not flood.  When the apparent volume of liquid
      reaches  0.5 ml, remove the K-D apparatus and allow it to drain  and
      cool for at least 10 minutes.

            7.1.2.5     Remove the micro-Snyder column and rinse the flask
      and  its lower  joint  into the  concentrator  tube with  0.2 mL of
      hexane.  Adjust the extract volume  to 2.0 ml and proceed with either
      Method 3610 or 3620.

7.2   Cleanup

      7.2.1 Proceed with   Method 3620,   using  the 2 ml  hexane extracts
obtained from  Section 7.1.2.5.

      7.2.2 Following cleanup, the extracts should be analyzed by GC, as
described in the  previous paragraphs and in Method 8000.

7.3   Gas Chromatography Conditions

      7.3.1 Table 1  summarizes  the  recommended operating conditions  for
the gas chromatograph.  This table includes retention times and MDLs that

                             8110 - 6                         Revision 0
                                                               July  1992

-------
      were  obtained  under  these  conditions.    Examples  of  the  parameter
      separations achieved by these  columns are shown in Figures  1 and 2.  Other
      packed columns, chromatographic conditions, or detectors  may be  used if
      the requirements of Section 8.2 are met.  Capillary (open-tubular)  columns
      may also  be  used  if the  relative standard deviations of responses  for
      replicate  injections  are  demonstrated  to  be  less  than   6%  and  the
      requirements of Section 8.2 are met.

      7.4   Calibration -  Refer to Method 8000 for proper calibration techniques.
Use Table 1  and  especially Table 2 for guidance on selecting the  lowest point on
the calibration curve.

            7.4.1 The procedure  for internal  or external  calibration  may be
      used.   Refer  to Method 8000 for a description of each of these procedures.

            7.4.2 If  cleanup  is  performed on  the samples, the  analyst should
      process a series  of  standards  through  the cleanup  procedure and  then
      analyze the  samples  by  GC.   This will confirm elution patterns  and  the
      absence of interferents from the reagents.

      7.5   Gas chromatographic analysis

            7.5.1 Refer to Method 8000.   If the internal  standard  calibration
      technique is used, add 10 nL of internal standard to the  sample prior to
      injection.

            7.5.2 Method 8000 provides  instructions on the  analysis sequence,
      appropriate  dilutions,  establishing daily  retention  time windows,  and
      identification criteria.   Include a mid-concentration check standard after
      each group of  10 samples in the analysis sequence.

            7.5.3 Examples of GC/HECD  chromatograms for haloethers  are shown in
      Figures 1 and  2.

            7.5.4 Record the sample  volume injected and  the resulting peak sizes
      (in area units or peak heights).

            7.5.5 Using either  the  internal or  external  calibration procedure
      (Method 8000),  determine the identity and quantity of each  analyte peak in
      the sample chromatogram.  See Method 8000  for  calculation  equations.

            7.5.6 If  peak detection and  identification  are  prevented  due to
      interferences,  the hexane  extract  may undergo cleanup using either Method
      3610 or 3620.
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was performed, follow the QC
in Method 3600 and in the specific cleanup method.
                                   8110 - 7                         Revision 0
                                                                     July 1992

-------
      8.2   Procedures to check the GC system operation are found in Method 8000,
Section 8.6.

            8.2.1 The quality control  (QC) reference sample concentrate (Method
      8000, Section 8.6) should contain each analyte of interest at 20 mg/L.

            8.2.2 Table  1   indicates   the  recommended  operating  conditions,
      retention  times,  and MDLs  that were  obtained  under  these  conditions.
      Table 2 gives method  accuracy and precision for the analytes of interest.
      The  contents  of both Tables  should be  used  to evaluate a laboratory's
      ability to perform and generate acceptable data by this method.

      8.3   Calculate surrogate standard recovery  on  all  samples,  blanks,  and
spikes.   Determine if  the recovery  is within  limits  (limits  established by
performing QC procedures outlined in Method 8000, Section 8.10).

            8.3.1 If recovery is not within limits, the following is required.

            •     Check  to  be  sure that  there are no errors in calculations,
                  surrogate  solutions and  internal  standards.   Also,  check
                  instrument performance.

            •     Recalculate the data and/or reanalyze the  extract if any of
                  the above checks reveal  a problem.

            •     Reextract and reanalyze  the sample  if  none  of the  above are a
                  problem or flag the data as "estimated concentration."


9.0   METHOD PERFORMANCE

      9.1   This method  has  been  tested for  linearity of recovery from spiked
organic-free reagent  water  and  has  been demonstrated  to be applicable for the
concentration range from 4 x MDL to 1000 x MDL.

      9.2   In  a single laboratory  (Monsanto  Research Center),  using spiked
wastewater samples, the  average recoveries presented in Table  2 were obtained.
Each spiked sample was analyzed in triplicate on three  separate occasions. The
standard deviation of the percent recovery is also included  in Table 2.


10.0  REFERENCES

1.    Fed. Regist. 1984, 49, 43234; October 26.

2.    Mills, P.A. "Variation of Florisil Activity:  Simple Method for Measuring
      Absorbent Capacity and Its Use in Standardizing  Florisil Columns"; Journal
      of the Association of Official  Analytical  Chemists 1968,  51_,  29.

3.    Handbook   of   Analytical   Quality  Control  in   Water  and  Wastewater
      Laboratories; U.S. Environmental Protection Agency. Office of Research and
      Development.   Environmental   Monitoring   and  Support   Laboratory.   ORD
      Publication  Offices  of  Center  for Environmental  Research Information:
      Cincinnati, OH,  1979;  EPA-600/4-79-019.

                                    8110 - 8                         Revision  0
                                                                     July  1992

-------
4.    Methods  for  Chemical  Analysis of  Water and Wastes;  U.S.  Environmental
      Protection Agency.  Office  of Research  and Development.  Environmental
      Monitoring and Support Laboratory. ORD Publication Offices of Center for
      Environmental Research Information:  Cincinnati, OH, 1983; EPA-600/4-79-
      020.

5.    Burke,  J.A.  "Gas Chromatography for  Pesticide Residue Analysis;  Some
      Practical Aspects";  Journal of  the  Association of  Official  Analytical
      Chemists 1965, 48, 1037.

6.    "EPA Method Validation Study 21 Methods 611  (Haloethers)," Report for EPA
      Contract 68-03-2633.

7.    "Determination of Haloethers in Industrial  and Municipal  Wastewaters";
      Report for EPA Contract 68-03-2633 (In preparation).
                                   8110 - 9                         Revision 0
                                                                     July 1992

-------
                                   TABLE 1.
            CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS
Analyte
  Retention Time
     (minutes)	
Column 1    Column 2
    Method
Detection Limit
    (M9/L)
Bis(2-chloroisopropyl) ether
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
4-Chlorophenyl phenyl ether
4-Bromophenyl phenyl ether
8.4
9.4
13.1
19.4
21.2
9.7
9.1
10.0
15.0
16.2
0.8
0.3
0.5
3.9
2.3
Column 1 conditions:
   Carrier gas (He) flow rate:
   Initial temperature:
   Temperature program:
   Final temperature:
  40 mL/min
  60°C> hold for 2 minutes
  60°C to 230°C at 8°C/min
  230°C, hold for 4 minutes
   Under these conditions the retention time for aldrin is 22.6 minutes.
Column 2 conditions:
   Carrier gas (He) flow rate:
   Initial temperature:
   Temperature program:
   Final temperature:
   40 mL/min
   150°C, hold for 4 minutes
   150°C to 310°C at 16°C/min
   310°C
   Under these conditions the retention time for aldrin  is  18.4 minutes.
                                   8110 - 10
                                       Revision  0
                                        July 1992

-------
                                   TABLE 2.
                    SINGLE OPERATOR ACCURACY AND PRECISION
                             Average   Standard    Spike    Number
                             Percent   Deviation   Range      of     Matrix
Analyte                      Recovery      %       (M9/L)  Analyses  Types

Bi s (2 -chl oroethoxyJmethane62       57313827       3
Bis(2-chloroethyl) ether         59       4.5         97        27       3
Bis(2-chloroisopropyl) ether     67       4.0         54        27       3
4-Bromophenyl phenyl ether       78       3.5         14        27       3
4-Chlorophenyl phenyl ether      73       4.5         30        27       3
                                   8110 -  11                         Revision 0
                                                                     July 1992

-------
               FIGURE  1.
   GAS CHROMATOGRAM OF HALOETHERS
Column: 3% SP-10OO on Suptleoport
Progrtm: 60°C. -2 mmuttt 8"/mmut» to 23O°C.
Detector: Hull »/»ctrolytic conductivity
 1

                     -c
                     5
                     i
                     5
   2  4   6   8  to  12  14  16  18  20  22 24
               8110  -  12
Revision 0
 July 1992

-------
                FIGURE 2.
    GAS CHROMATOGRAM OF HALOETHERS
Column: Tenex GC
Progrem: 150°C.-4 minutes 16*/minute to 310*C.
Detector: Hall electrolytic conductivity
                   12      16     20

             Retention time, minutes
24
                8110 - 13
             Revision  0
              July 1992

-------
                        METHOD 8110
           HALOETHERS BY  GAS  CHROMATOGRAPHY
                              Start
  7124 Perform
micro-K D procedure
   using hexane,
proceed with Method
   3610 or 3620
                          711  Choose
                           appr oprlate
                           extraction
                            pr ocedure
                          712  Perform
                        solvent  exchange
                          using  hexane
                         7123  Adjust
                        extract volume and
                          proceed with
                        ana lysis  or store
                         in appropriate
                             manne r
                         7 3 1  Refer to
                           Table  1 for
                           r ecommended
                            o pe ra 11ng
                        conditions for the
                               CC
4 Refer to Method
8000  for proper
  ca 1 ibra tion
  techniques
751 Refer to
Method 8000 for
guidance on CC
anal ys is


7 5 4 Record sample
volume injected and
resul ting

peak size

755 Perform
appr opr la te
calculations (refer
to Method 8000)


     Stop
                          8110  -  14
                   Revision 0
                    July  1992

-------
                                 METHOD 8120A

                CHLORINATED HYDROCARBONS BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method  8120  is  used to  determine  the  concentration  of certain
chlorinated hydrocarbons.   The following compounds can  be  determined by this
method:
                                            Appropriate Preparation Techniques

Compounds                         CAS No"     3510     3520  3540/   3550   3580
                                                             3541
2-Chloronaphthalene
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachl orobutadi ene
Hexachl orocycl ohexane
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachlorohexane
Tetrachl orobenzenes
1, 2, 4-Tri chl orobenzene
91-58-7
95-50-1
541-73-1
106-46-7
118-74-1
87-68-3
608-73-1
77-47-4
67-72-1

--
120-82-1
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
   a  Chemical Abstract Services Registry Number.
   x  Greater than 70 percent recovery by this technique
   ND Not determined.

      1.2   Table 1 indicates compounds that may be determined by this method and
lists the method detection limit for each compound in organic-free reagent water.
Table 2 lists the estimated quantitation limit (EQL) for other matrices.


2.0   SUMMARY OF METHOD

      2.1   Method 8120 provides gas chromatographic conditions for the detection
of ppb concentrations of certain chlorinated hydrocarbons.  Prior to use of this
method, appropriate  sample extraction  techniques must  be used.   Both  neat and
diluted organic liquids (Method 3580,  Waste Dilution) may be analyzed by direct
injection.    A  2 to 5 juL aliquot  of the  extract  is  injected  into a  gas
chromatograph (GC),  and compounds in the GC effluent are detected by an electron
capture detector (ECD).
                                  8120A  -  1                         Revision 1
                                                                September 1994

-------
      2.2   If interferences are encountered in the  analysis,  Method 8120 may
also be performed on extracts that have undergone cleanup using Method 3620.

3.0   INTERFERENCES

      3.1   Refer to Methods 3500,  3600,  and  8000.

      3.2   Solvents,  reagents,  glassware, and other sample processing hardware
may yield discrete artifacts and/or elevated baselines causing misinterpretation
of gas chromatograms.  All  of these  materials must  be  demonstrated to be free
from interferences, under the conditions  of  the  analysis,  by analyzing method
blanks.    Specific selection  of  reagents  and  purification  of   solvents  by
distillation in all glass systems may be required.

      3.3   Interferences coextracted from samples  will  vary considerably from
source to  source, depending upon  the  waste  being  sampled.   Although general
cleanup techniques  are recommended as  part of this  method,  unique samples may
require additional cleanup.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1  Gas  chromatograph  - Analytical  system  complete  with  gas
      chromatograph  suitable   for   on-column  injections   and  all  required
      accessories,  including detectors, column supplies,  recorder, gases, and
      syringes.  A data system for measuring peak areas and/or peak heights is
      recommended.

            4.1.2  Columns

                   4.1.2.1     Column 1  -  1.8 m  x 2 mm  ID  glass  column packed
            with 1% SP-1000  on  Supelcoport (100/120  mesh)  or equivalent.

                   4.1.2.2     Column 2  -  1.8 m  x 2 mm  ID  glass  column packed
            with  1.5%  OV-1/2.4%  OV-225  on  Supelcoport  (80/100  mesh)  or
            equivalent.
      4.2
4.1.3  Detector - Electron capture (ECD).

Kuderna-Danish (K-D) apparatus
            4.2.1  Concentrator tube - 10 ml,  graduated  (Kontes K-570050-1025 or
      equivalent).  A  ground  glass stopper is used  to prevent  evaporation of
      extracts

            4.2.2  Evaporation  flask   -   500  ml   (Kontes   K-570001-500  or
      equivalent).   Attach   to  concentrator  tube  with  springs,  clamps  or
      equivalent.

            4.2.3  Snyder column - Three  ball macro (Kontes  K-503000-0121 or
      equivalent).
                                   8120A -  2
                                                        Revision 1
                                                    September 1994

-------
            4.2.4  Snyder column  -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.2.5  Springs  -   1/2  inch  (Kontes K-662750  or  equivalent).

      4.3   Boiling chips - Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

      4.4   Water  bath  -  Heated,  with  concentric  ring   cover,  capable  of
temperature control (+ 5°C).   The  bath  should  be used in a  hood.

      4.5   Volumetric flasks -  10,  50,  and  100 ml, with ground glass stoppers.

      4.6   Microsyringe  -  10 juL.

      4.7   Syringe -  5 ml.

      4.8   Vials - Glass,  2, 10,  and  20  ml capacity with  Teflon lined screw-
caps or crimp tops.


5.0   REAGENTS

      5.1   Reagent grade inorganic  chemicals shall be used  in  all tests. Unless
otherwise  indicated,  it  is  intended  that  all  reagents shall conform  to the
specifications of the Committee  on Analytical Reagents of the American Chemical
Society, where  such  specifications  are available.   Other  grades  may be used,
provided it is first ascertained that the reagent is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent water.  All references  to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Solvents

            5.3.1  Hexane, C6H14.   Pesticide  quality or equivalent.

            5.3.2  Acetone, CH3COCH3.  Pesticide quality or equivalent.

            5.3.3  Isooctane,  C8H18.   Pesticide  quality or equivalent.

      5.4   Stock standard  solutions

            5.4.1  Prepare  stock standard  solutions at a concentration of 1000
      mg/L by dissolving  0.0100 g  of assayed reference material in isooctane or
      hexane and diluting to  volume  in a 10 ml volumetric flask.  Larger volumes
      can be used at the convenience of the analyst.   When compound  purity is
      assayed to be 96% or greater, the weight can  be  used without correction to
      calculate the concentration  of the stock standard.  Commercially prepared
      stock standards  can be  used  at any concentration if they are certified by
      the manufacturer or by  an  independent  source.
                                  8120A  - 3                         Revision 1
                                                                September 1994

-------
            5.4.2  Transfer  the  stock standard solutions into vials with Teflon
      lined screw  caps  or crimp tops.   Store  at 4°C and protect  from light.
      Stock standards should be checked frequently for signs of degradation or
      evaporation,  especially just prior to preparing calibration standards.

            5.4.3  Stock standard solutions must be replaced after one year, or
      sooner if comparison with check standards  indicates  a  problem.

      5.5   Calibration  standards  - Calibration  standards  at a minimum of five
concentrations should be prepared through dilution of the stock standards with
isooctane or hexane.  One of the  concentrations  should be  at a concentration
near, but above,  the method detection limit.  The  remaining concentrations should
correspond to  the  expected range  of  concentrations  found  in  real  samples or
should define  the  working  range  of  the  GC.   Calibration  solutions  must be
replaced after  six months, or sooner if comparison with check  standards indicates
a problem.

      5.6   Internal  standards (if internal  standard  calibration is used) - To
use this approach,  the analyst must select one or more internal standards that
are similar in analytical behavior  to  the  compounds  of  interest.   The analyst
must further demonstrate  that the  measurement of  the internal  standard is not
affected by method or matrix  interferences.   Because of these limitations, no
internal  standard can be suggested that is applicable to all samples.

            5.6.1  Prepare  calibration  standards   at   a  minimum   of   five
      concentrations for each analyte of interest as  described in Sec.  5.5.

            5.6.2  To each calibration  standard,  add  a known constant amount of
      one or more  internal  standards,  and dilute to  volume with  isooctane or
      hexane.

            5.6.3  Analyze each  calibration  standard  according  to Sec.  7.0.

      5.7   Surrogate standards  - The analyst should  monitor the performance of
the extraction, cleanup  (when used), and analytical system  and the effectiveness
of  the  method  in  dealing  with each  sample matrix  by spiking each  sample,
standard, and organic-free reagent water blank with one  or two surrogates (e.g.
chlorinated hydrocarbons that are  not  expected to  be  in the  sample)  recommended
to encompass the range of the temperature program used in this method.  Method
3500  details   instructions  on  the  preparation  of   base/neutral  surrogates.
Deuterated  analogs of  analytes  should not  be  used as  surrogates  for  gas
chromatographic  analysis due to coelution problems.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material to this chapter, Organic Analytes, Sec.
4.1.

      6.2   Extracts must be stored under refrigeration  and analyzed within 40
days of extraction.
                                   8120A -  4                         Revision 1
                                                                September 1994

-------
7.0   PROCEDURE

      7.1    Extraction

            7.1.1  Refer to Chapter Two for guidance on choosing the appropriate
      extraction procedure.    In  general, water  samples are  extracted at  a
      neutral,  or as is, pH with methylene chloride, using either Method 3510 or
      3520.  Solid samples are extracted using either Methods 3540/3541 or 3550.

            7.1.2  Prior to gas chromatographic analysis, the extraction solvent
      must be exchanged to  hexane.   The exchange is performed during  the  K-D
      procedures listed  in  all of  the extraction methods.   The  exchange  is
      performed as  follows.

                  7.1.2.1     Following K-D of the methylene  chloride  extract
            to  1 ml using  the macro  Snyder column,  allow the  apparatus  to cool
            and  drain for  at  least 10  minutes.

                  7.1.2.2     Momentarily remove the Snyder column,  add 50  mi
            of hexane, a new boiling  chip,  and  reattach the macro Snyder column.
            Concentrate  the  extract  using 1  ml of hexane to prewet the Snyder
            column.   Place  the K-D  apparatus  on  the water  bath  so that  the
            concentrator  tube is partially immersed in the hot water.   Adjust
            the  vertical position of the apparatus  and the water temperature,  as
            required, to complete concentration in 5-10 minutes.  At the proper
            rate of distillation the balls of the column will  actively chatter,
            but  the chambers will not flood.  When the apparent volume of liquid
            reaches 1 ml,  remove the  K-D  apparatus  and allow it to  drain  and
            cool  for  at  least  10  minutes.    The  extract  will  be  handled
            differently  at this point, depending on whether or not cleanup  is
            needed.   If  cleanup is not required, proceed to Sec. 7.1.2.3.   If
            cleanup is needed,  proceed  to  Sec.  7.1.2.4.

                  7.1.2.3     If cleanup of the extract is not required, remove
            the  Snyder column  and  rinse the flask  and  its lower joint into  the
            concentrator  tube  with  1-2  ml  of  hexane.   A  5 ml  syringe  is
            recommended  for  this  operation.   Adjust  the  extract volume  to
            10.0 ml.  Stopper the concentrator tube and store refrigerated at 4°C
            if  further  processing will not be  performed immediately.   If  the
            extract  will  be   stored,   longer  than  two  days,  it   should   be
            transferred  to a vial  with a Teflon lined screw cap or crimp top.
            Proceed with gas  chromatographic analysis.

                  7.1.2.4     If cleanup  of the extract is required, remove  the
            Snyder  column  and  rinse  the  flask  and  its  lower joint into  the
            concentrator tube  with a minimum  amount of  hexane.  A 5 ml  syringe
            is recommended for  this operation.  Add a clean  boiling  chip to  the
            concentrator tube and attach a two  ball micro Snyder column.   Prewet
            the  column by  adding about  0.5 ml  of hexane  to the top.   Place the
            micro K-D apparatus on  the water bath (80°C) so that the concentrator
            tube is partially immersed in the hot water.  Adjust the vertical
            position of the apparatus and  the water temperature, as  required,  to
            complete  concentration in  5-10  minutes.    At the  proper rate  of


                                  8120A - 5                         Revision  1
                                                               September 1994

-------
            distillation the  balls of the column will actively chatter, but the
            chambers will not flood.  When the apparent  volume of liquid reaches
            0.5 ml,  remove  the  K-D apparatus  and allow  it to drain and cool for
            at  least 10  minutes.

                  7.1.2.5     Remove the micro Snyder column and rinse the flask
            and its  lower  joint  into  the  concentrator  tube with  0.2 ml  of
            hexane.   Adjust the extract volume to 2.0 ml and proceed with Method
            3620.

      7.2   Gas chromatographic conditions  (Recommended)

            7.2.1  Column 1

            Carrier  gas  (5% methane/95% argon) flow rate  =  25 mL/min
            Column temperature  =    65°C  isothermal, unless otherwise specified
                                    (see Table 1).

            7.2.2  Column 2

            Carrier  gas  (5% methane/95% argon) flow rate  =  25 mL/min
            Column temperature  =    75°C  isothermal, unless otherwise specified
                                    (see Table 1).

      7.3   Calibration  - Refer to Method 8000 for proper calibration techniques.
Use Table 1  and especially Table 2 for guidance on selecting the lowest point on
the calibration curve.

            7.3.1  The procedure for internal or  external  calibration  may  be
      used.   Refer to Method 8000 for a  description of each of these procedures.

            7.3.2  If cleanup is performed on the samples,  the  analyst should
      process  a  series  of standards  through the cleanup  procedure  and  then
      analyze the samples by GC.  This will  validate elution  patterns and the
      absence of interferents from the  reagents.

      7.4   Gas chromatographic analysis

            7.4.1  Refer to Method  8000.   If the  internal standard calibration
      technique is  used, add  10 /xL of internal standard to the sample prior to
      injecting.

            7.4.2  Method 8000  provides instructions  on the analysis sequence,
      appropriate dilutions,  establishing  daily retention  time  windows,  and
      identification criteria.   Include a mid-concentration standard after each
      group  of 10 samples in  the analysis sequence.

            7.4.3  Examples  of  GC/ECD  chromatograms  for certain  chlorinated
      hydrocarbons  are shown  in Figures 1 and 2.

            7.4.4  Record the sample volume  injected and the  resulting peak sizes
      (in area units or peak  heights).
                                   8120A -  6                         Revision 1
                                                                September 1994

-------
            7.4.5  Using  either the internal  or external  calibration procedure
      (Method 8000),  determine the  identity and quantity of each component peak
      in the  sample  chromatogram  which corresponds to the  compounds  used for
      calibration purposes.  See Method 8000  for calculation equations.

            7.4.6  If peak detection  and  identification  are prevented  due to
      interferences,  the hexane extract may undergo cleanup using Method 3620.

      7.5   Cleanup:  If required, the samples  may be cleaned up using the Methods
presented in Chapter 4.

            7.5.1  Proceed  with Method 3620   using  the 2  ml  hexane  extracts
      obtained from Sec. 7.1.2.5.

            7.5.2  Following  cleanup,  the extracts should be analyzed by GC, as
      described in the previous paragraphs and in Method  8000.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter  One   for specific  quality  control  procedures.
Quality control to validate sample extraction is covered  in Method 3500 and in
the extraction method utilized.  If extract cleanup was  performed, follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Procedures to check the GC system  operation are  found in Method 8000.

            8.2.1  The quality control check  sample concentrate (Method 8000)
      should contain  each  parameter of interest at the following concentrations
      in acetone:  hexachloro-substituted  hydrocarbon,  10  mg/L;  and  any other
      chlorinated hydrocarbon, 100 mg/L.

            8.2.2  Table 3  indicates  the calibration and QC acceptance criteria
      for this method.  Table 4 gives method accuracy and precision as functions
      of concentration  for the analytes  of   interest.   The contents of  both
      Tables should be used to  evaluate a laboratory's ability to  perform and
      generate acceptable data by  this method.

      8.3   Calculate surrogate standard recovery on all  samples,  blanks,  and
spikes.   Determine if  the  recovery  is within  limits  (limits  established by
performing QC procedures outlined  in Method 8000).

            8.3.1  If recovery is not within limits, the  following procedures are
      required.

                   •     Check  to be  sure  there  are no  errors in calculations,
                        surrogate  solutions   and  internal  standards.    Also,
                        check  instrument performance.

                   •     Recalculate  the data and/or  reanalyze  the  extract if
                        any of the above checks reveal  a problem.
                                  8120A  - 7                         Revision 1
                                                                September 1994

-------
                        Reextract and reanalyze the sample if none of the above
                        are   a   problem  or  flag  the   data  as   "estimated
                        concentration".
9.0   METHOD PERFORMANCE

      9.1   The method  was  tested  by 20 laboratories using organic-free reagent
water, drinking water,  surface water, and three industrial wastewaters spiked at
six concentrations over the range 1.0 to 356 M9/L.  Single operator precision,
overall precision, and method accuracy were found  to be directly related to the
concentration of the  parameter and essentially independent of the sample matrix.
Linear equations to describe these relationships for a  flame  ionization detector
are presented in Table 4.

      9.2   The accuracy  and precision obtained will be determined by the sample
matrix, sample preparation technique, and calibration procedures used.


10.0  REFERENCES

1.    "Development and  Application of Test Procedures for  Specific Organic Toxic
      Substances in  Wastewaters.   Category 3  - Chlorinated Hydrocarbons,  and
      Category 8 - Phenols,"  Report for EPA Contract 68-03-2625.

2.    Burke,  J.A.  "Gas  Chromatography  for  Pesticide Residue Analysis;  Some
      Practical Aspects,"  Journal  of  the  Association of  Official  Analytical
      Chemists, 48, 1037, 1965.

3.    "EPA Method Validation  Study 22, Method 612   (Chlorinated Hydrocarbons),"
      Report for EPA Contract 68-03-2625.

4.    "Method  Performance  for   Hexachlorocyclopentadiene   by  Method  612,"
      Memorandum  from   R.  Slater,   U.S.   Environmental   Protection  Agency,
      Environmental Monitoring and Support Laboratory, Cincinnati,  Ohio 45268,
      December 7, 1983.

5.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
      Analysis of Pollutants  Under the Clean Water Act;  Final Rule and Interim
      Final Rule and Proposed Rule," October 26,  1984.

6.    "Determination of  Chlorinated  Hydrocarbons   in  Industrial  and Municipal
      Wastewaters," Report for EPA Contract 68-03-2625.
                                   8120A -  8                        Revision 1
                                                                September 1994

-------
                                   TABLE 1.
                GAS CHROMATOGRAPHY OF CHLORINATED HYDROCARBONS
Compound
Retention time (min)

Col. 1       Col. 2
   ND = Not determined.

   3150°C column temperature,

   b!65°C column temperature.

   C100°C column temperature.
  Method
 Detection
limit (/ig/L)
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachlorobutadiene
Hexachl orocycl ohexane
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachlorohexane
Tetrachl orobenzenes
1,2,4-Trichlorobenzene
2.7a
6.6
4.5
5.2
5.6"
7.7

ND
4.9


15.5
3.6"
9.3
6.8
7.6
10. lb
20.0

16.5°
8.3


22.3
0.94
1.14
1.19
1.34
0.05
0.34

0.40
0.03


0.05
                                  8120A  -  9
                                    Revision 1
                                September 1994

-------
                                TABLE 2.
                 DETERMINATION  OF  ESTIMATED  QUANTITATION
                   LIMITS (EQL) FOR VARIOUS MATRICES'
Matrix                                                             Factor
Ground water                                                            10
Low-concentration soil by ultrasonic extraction with GPC cleanup       670
High-concentration soil and sludges by ultrasonic extraction        10,000
Non-water miscible waste                                           100,000
   a  EQL = [Method detection limit  (see  Table  1)]  X [Factor found in this
      table].   For non-aqueous  samples,  the  factor is on a wet weight basis.
      Sample EQLs  are  highly matrix  dependent.   The EQLs  to be determined
      herein are provided for guidance and may not always be achievable.
                               8120A - 10                        Revision 1
                                                             September 1994

-------
                        TABLE 3.
                 QC ACCEPTANCE  CRITERIA8


Parameter
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachl orobutadi ene
Hexachl orocycl opentadi ene
Hexachl oroethane
1,2, 4-Tri chl orobenzene
Test
cone.
Ug/L)
100
100
100
100
10
10
10
10
100
Limit Range
for s for x
(M9/L) (M9/L)
37.3 29.5-126.9
28.3 23.5-145.1
26.4 7.2-138.6
20.8 22.7-126.9
2.4 2.6-14.8
2.2 D-12.7
2.5 D-10.4
3.3 2.4-12.3
31.6 20.2-133.7
Range
P> P.
(%)
9-148
9-160
D-150
13-137
15-159
D-139
D-lll
8-139
5-149
s = Standard deviation of four recovery measurements, in jug/L.
x = Average recovery
P,PS = Percent recovery
D = Detected; result
a Criteria from 40
for four recovery
measured.
measurements, in fj.g/1.



must be greater than zero.
CFR Part 136 for
Method 612. These criteria are
based directly upon the method performance data in Table 4.  Where
necessary, the  limits  for recovery have  been  broadened to assure
applicability of the limits  to  concentrations  below those used to
develop Table 4.
                      8120A - 11                        Revision 1
                                                    September 1994

-------
                                   TABLE 4.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION8


Parameter
Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobutadiene
Hexachl orocycl opentadi enea
Hexachloroethane
1 , 2 , 4-Tri chl orobenzene
Accuracy, as
recovery, x'
(M9/L)
0.75C+3.21
0.85C-0.70
0.72C+0.87
0.72C+2.80
0.87C-0.02
0.61C+0.03
0.47C
0.74C-0.02
0.76C+0.98
Single analyst
precision, s/
(M9/L)
0.28X-1.17
0.22X-2.95
0.21X-1.03
0.16X-0.48
0.14X+0.07
0.18X+0.08
0.24x
0.23X+0.07
0.23X-0.44
Overall
precision,
S' (M9/L)
0.38X-1.39
0.41X-3.92
0.49X-3.98
0.35X-0.57
0.36X-0.19
0.53X-0.12
0.50x
0.36X-0.00
0.40X-1.37
x'


V


S'


C

x
Expected  recovery  for  one  or  more  measurements   of   a  sample
containing a concentration of C, in M9/L.

Expected single  analyst  standard deviation of  measurements at an
average concentration of x, in
Expected interlaboratory standajrd  deviation  of measurements at an
average concentration found of x, in M9/L.

True value for the concentration, in M9/L-

Average recovery  found for  measurements  of  samples  containing  a
concentration of C, in /xg/L.

Estimates based upon the performance in a single laboratory.
                                  8120A - 12
                                                         Revision  1
                                                     September  1994

-------
                                   FIGURE  1
              Column:       1.5%  OV-1  +  1.5% OV-225 on Ga» Chrom. Q
              Temperature:  7 5 ° C
              Detector:     Electron Capture
                                             I
                  tt	•
                                   tit    i	  i
                  4        •       12        16
                    RETENTION TIM! (MINUTIS)
20
Gas chromatagram of chlorinated hydrocarbons  (high molecular weight compounds).
                                  8120A -  13
                Revision  1
           September  1994

-------
                                     FIGURE 2
                           Column:      1.5% OV-1 +1.5* OV-225 on Gas Chrom.  Q
                           Temperature:  16 0 °C
                           Detector:    Electron Capture
                   0       4        •       12      16
                          RETENTION TIME (MINUTES)
Gas chromatagram of chlorinated hydrocarbons (low molecular weight compounds)
                                     8120A  -  14
    Revision 1
September 1994

-------
                                   METHOD  8120A
              CHLORINATED  HYDROCARBONS BY  GAS  CHROMATOGRAPHY
 (    Start    J
  7.1.1 Choose
   appropriate
    extraction
  procedure (see
   Chapter 2).
 7.1.2 Exchange
extraction so/vent
 to hexane during
 K-D procedures.
   7.2 Set gas
 chromatography
   conditions.
7.3 Refer to Method
  8000 for proper
    calibration
    techniques.
   7.3.2 Is
   cleanup
 necessary?
 7.3.2 Process a
series of standards
 through cleanup
procedure; analyze
     by GC.
7.4 Perform GC
 analysis (see
Method 8000).
                                       7.4.5
                                   Is identification
                                     & detection
                                     prevented by
                                    interferences?
                            7.5.1 Cleanup using
                               Method 3620.
                                    8120A -  15
                                           Revision  1
                                      September  1994

-------
                                  METHOD 8121

  CHLORINATED HYDROCARBONS BY GAS CHROMATOGRAPHY:  CAPILLARY COLUMN TECHNIQUE


1.0   SCOPE AND APPLICATION

      1.1   Method 8121 describes the determination of chlorinated hydrocarbons
in extracts prepared from environmental samples and RCRA wastes.   It describes
wide-bore open-tubular, capillary column gas chromatography procedures using both
single column/single  detector and dual-column/dual-detector approaches.   The
following compounds can be determined by this method:


      Compound Name                           CAS  Registry No."

      Benzal chloride98-87-3
      Benzotrichloride                           98-07-7
      Benzyl chloride                           100-44-7
      2-Chloronaphthalene                        91-58-7
      1,2-Dichlorobenzene                        95-50-1
      1,3-Dichlorobenzene                       541-73-1
      1,4-Dichlorobenzene                       106-46-1
      Hexachlorobenzene                         118-74-1
      Hexachlorobutadiene                        87-68-3
      a-Hexachlorocyclohexane (a-BHC)           319-84-6
      0-Hexachlorocyclohexane (/3-BHC)           319-85-7
      7-Hexachlorocyclohexane (7-BHC)            58-89-9
      6-Hexachlorocyclohexane (5-BHC)           319-86-8
      Hexachlorocyclopentadiene                  77-47-4
      Hexachloroethane                           67-72-1
      Pentachlorobenzene                        608-93-5
      1,2,3,4-Tetrachlorobenzene                634-66-2
      1,2,4,5-Tetrachlorobenzene                 95-94-2
      1,2,3,5-Tetrachlorobenzene                634-90-2
      1,2,4-Trichlorobenzene                    120-82-1
      1,2,3-Trichlorobenzene                     87-61-6
      1,3,5-Trichlorobenzene                    108-70-3

      a   Chemical  Abstract Services  Registry  Number.

      1.2   The dual-column/dual-detector  approach  involves  the  use of  two
30 m x 0.53 mm ID  fused-sil ica open-tubular columns of different polarities, thus
different selectivities towards  the target compounds.  The columns are connected
to an injection tee and two  identical  detectors.   When  compared  to the packed
columns,   the  megabore   fused-sil ica  open-tubular  columns  offer  improved
resolution, better selectivity,  increased sensitivity,  and faster analysis.

      1.3   Table  1 lists  method detection limits (MDL) for each  compound in an
organic-free reagent water matrix.   The  MDLs for the compounds  of a  specific
sample may differ  from those listed  in  Table  1 because  they are  dependent upon
                                   8121 - 1                         Revision 0
                                                                September 1994

-------
the nature of interferences in the sample matrix.   Table 2 lists the estimated
quantitation limits (EQL) for other matrices.

      1.4   Table 3 lists the compounds that have been determined by this method
and their retention times using the single column technique.  Table 4 lists dual
column/dual detector retention time  data.   Figures  1 and 2  are chromatograms
showing the single  column technique.  Figure 3 shows a chromatogram of the target
analytes eluted from a pair of DB-5/DB-1701  columns and  detected with electron
capture detectors  (ECD) under the prescribed GC conditions listed in Table 2.

      1.5   This method  is  restricted to use  by  or under the  supervision  of
analysts experienced in the use of a gas  chromatograph and in the interpretation
of gas chromatograms.


2.0   SUMMARY OF METHOD

      2.1   Method 8121 provides gas chromatographic conditions for the detection
of ppb  concentrations  of chlorinated  hydrocarbons  in  water  and soil  or ppm
concentrations in waste samples.  Prior to use of this method, appropriate sample
extraction techniques must be used for environmental samples (refer to Chapt. 2).
Both neat and diluted  organic liquids (Method  3580)  may be  analyzed by direct
injection.  Spiked  samples are used  to  verify  the  applicability of the chosen
extraction technique to each new sample type.   Analysis  is accomplished by gas
chromatography utilizing  an instrument equipped with wide bore capillary columns
and single or dual  electron capture detectors.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000.

      3.2   The  electron capture  detector  responds to  all  electronegative
compounds. Therefore, interferences are possible by  other halogenated compounds,
as well  as phthalates  and  other  oxygenated  compounds,  and,  organonitrogen,
organosulfur and organophosphorus compounds.  Second column confirmation or GC/MS
confirmation  are  necessary  to  ensure  proper  analyte  identification  unless
previous characterization of the sample source will ensure proper identification.

      3.3   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.  To reduce carryover, the
syringe  used  for injection  must be  rinsed  out between samples  with  solvent.
Whenever  an extract  concentration  exceeds  that  of the highest  calibration
standard, it should be followed  by  the analysis  of  a  solvent blank to check for
cross-contamination.   Additional  solvent  blanks interspersed with  the sample
extracts should be  considered whenever the analysis  of a  solvent blank indicates
cross-contamination problems.

      3.4   Phthalate esters,  if present in a sample, will interfere only with
the BHC  isomers because  they  elute  in Fraction  2 of  the  Florisil  procedure
described  in  Method 3620.  The  presence of  phthalate  esters can  usually  be
minimized  by  avoiding contact  with  any  plastic   materials  and by  following
standard decontamination  procedures of reagents and glassware.

                                   8121  - 2                         Revision 0
                                                                September 1994

-------
      3.5   The presence of elemental sulfur will result in large peaks, and can
often mask the region of compounds eluting after 1,2,4,5-tetrachlorobenzene.  The
tetrabutylammonium  (TBA)-sulfite  procedure (Method  3660)  works well  for  the
removal of elemental sulfur.

      3.6   In  certain cases  some compounds  coelute  on  either  one  or both
columns.   In  these cases  the compounds must  be  reported as  coeluting.   The
mixture can be reanalyzed by GC/MS techniques, see Sec. 8.7 and Method 8270.

            3.6.1 Using  the  dual  column  system  of  analysis the  following
      compounds coeluted:

            DB-5        1,4-dichlorobenzene/benzyl chloride
                        l,2,3,5-tetrachlorobenzene/l,2,4,5-tetrachlorobenzene
                        l,2,3,4-tetrachlorobenzene/2-chloronaphthalene

            DB-1701     benzyl chloride/1,2-dichlorobenzene/hexachloroethane
                        benzal chloride/1,2,4-trichlorobenzene/
                        hexachlorobutadiene

            Some of the injections showed a separation of 1,2,4-trichlorobenzene
      from the other two compounds, however, this is not always the case,  so the
      compounds are listed as coeluting.

      3.7   Solvents, reagents, glassware,  and other  sample processing hardware
may yield discrete artifacts and/or elevated baselines causing misinterpretation
of gas chromatograms.   All  these materials  must be demonstrated to be free from
interferences under the conditions of the analysis,  by analyzing reagent blanks.


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph: An  analytical system complete  with   a  gas
chromatograph suitable for on-column  and  split-splitless injection,  and  all
required  accessories,  including syringes,  analytical  columns, gases,  and  two
electron  capture detectors. A data system for measuring  peak areas,  and dual
display of chromatograms is recommended.  A GC equipped with a single GC column
and detector are acceptable,  however,  second  column  confirmation  is obviously
more time consuming.   Following are  the single and dual  column configurations
used for developing  the retention time data  presented  in the method.  The columns
listed  in the  dual  column  configuration may  also  be  used for single column
analysis.

            4.1.1 Single Column Analysis:

                  4.1.1.1     Column  1  -  30  m  x  0.53  mm  ID  fused-silica
            capillary  column   chemically  bonded with  trifluoropropyl  methyl
            silicone (DB-210 or equivalent).

                  4.1.1.2     Column  2  -  30  m  x  0.53  mm   ID  fused-silica
            capillary column chemically bonded with polyethylene glycol (DB-WAX
            or equivalent).


                                   8121 - 3                         Revision 0
                                                                September 1994

-------
            4.1.2 Dual  Column Analysis:

                  4.1.2.1     Column  1  -  30  m  x  0.53  mm  ID  fused-silica
            open-tubular  column,  crosslinked  and chemically  bonded with  95
            percent dimethyl and 5 percent diphenyl-polysiloxane (DB-5, RTx-5,
            SPB-5, or equivalent), 0.83 ptm or 1.5 /im film thickness.

                  4.1.2.2     Column  2  -  30  m  x  0.53  mm  ID  fused-silica
            open-tubular  column  crosslinked  and  chemically  bonded with  14
            percent  cyanopropylphenyl   and  86  percent  dimethyl-polysiloxane
            (DB-1701, RTX-1701,  or equivalent),  1.0 urn film thickness.

            4.1.3 Splitter: If the  splitter approach  to  dual  column injection
      is  chosen,  following  are  three suggested  splitters.    An  equivalent
      splitter is  acceptable.   See Sec.  7.5.1  for  a caution  on  the  use  of
      splitters.

                  4.1.3.1     Splitter  1  -  J&W Scientific  press-fit Y-shaped
            glass 3-way union splitter  (J&W Scientific, Catalog no. 705-0733).

                  4.1.3.2     Splitter  2 - Supelco 8  in.  glass  injection tee,
            deactivated (Supelco,  Catalog no.  2-3665M).

                  4,1.3.3     Splitter  3   -   Restek  Y-shaped   fused-silica
            connector (Restek,  Catalog no.  20405).

            4.1.4 Column rinsing kit (optional): Bonded-phase column rinse kit
      (J&W Scientific,  Catalog no. 430-3000 or equivalent).

            4.1.5 Microsyringes -  100  jiL,  50 /zL,  10 nl  (Hamilton 701  N  or
      equivalent), and 50 /iL (Blunted,  Hamilton 705SNR or equivalent).

            4.1.6 Balances - Analytical, 0.0001 g.

            4.1.7 Volumetric flasks, Class A - 10 ml to 1000 mL.
5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used in all  tests.  Unless
otherwise  indicated,  it is  intended  that all  reagents  shall  conform  to the
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where  such  specifications  are available.  Other grades  may be used,
provided it  is  first ascertained that the chemicals are  of sufficiently high
purity to permit their use without affecting the accuracy of the determinations.

      NOTE: Store  the  standard  solutions   (stock,  composite,   calibration,
            internal, and surrogate) at 4°C in Teflon-sealed containers in the
            dark.  All standard solutions must be replaced after six months or
            sooner if routine QC (Sec. 8) indicates a problem.
                                   8121 - 4
    Revision 0
September 1994

-------
      5.2   Solvents

            5.2.1 Hexane, C6H14  -  Pesticide quality or equivalent.

            5.2.2 Acetone, CH3COCH3 -  Pesticide quality  or equivalent.

            5.2.3 Isooctane, (CH3)3CCH2CH(CH3}2 - Pesticide quality or equivalent.

      5.3   Stock standard  solutions  (1000 mg/L):  Can  be  prepared  from pure
standard materials or can be purchased as certified solutions.

            5.3.1 Prepare stock standard solutions by accurately weighing about
      0.0100 g of pure compound.  Dissolve the compound in  isooctane or hexane
      and dilute to volume in a 10 ml volumetric flask.   If compound purity is
      96 percent  or  greater,  the weight  can  be used  without  correction  to
      calculate the concentration of the stock standard solution. Commercially
      prepared stock standard solutions  can be used at  any  concentration  if they
      are certified by the manufacturer or by an independent source.

            5.3.2 For those compounds which are not adequately soluble in hexane
      or isooctane,  mixtures of acetone and hexane are recommended.

      5.4   Composite  stock  standard:  Can  be  prepared  from  individual  stock
solutions.  For composite  stock standards  containing  less than 25 components,
take exactly 1 mi of each individual stock solution at 1000 mg/L, add solvent,
and mix the solutions  in  a 25 ml volumetric flask. For example, for a composite
containing 20  individual standards, the resulting concentration of each component
in the mixture, after the volume  is adjusted to 25  mL,  will be 40 mg/L.  This
composite solution can be further diluted to obtain the desired concentrations.

      5.5   Calibration  standards should  be  prepared  at  a  minimum  of five
concentrations by dilution  of the composite stock standard with isooctane or
hexane.    The concentrations   should  correspond  to  the  expected  range  of
concentrations found in real  samples and should  bracket the  linear range of the
detector.  A suggested list  of calibration  solution standards  is found in Table
7.

      5.6   Recommended  internal  standard: Make  a  solution  of 1000  mg/L  of
1,3,5-tribromobenzene.   (Two  other  internal  standards,  2,5-dibromotoluene and
alpha,alpha'-dibromo-m-xylene,  are  suggested  if matrix  interferences  are  a
problem.)  For spiking, dilute this solution to  50 ng//*L.   Use  a spiking volume
of 10 ij,L/mi of extract.   The spiking concentration of  the internal  standards
should be  kept constant  for all samples and calibration  standards.  Store the
internal standard spiking solutions at  4"C  in  Teflon-sealed containers in the
dark.

      5.7   Recommended surrogate  standards:   Monitor the  performance  of the
method using surrogate compounds.  Surrogate standards  are  added to all samples,
method blanks, matrix  spikes,  and calibration  standards.    Make  a  solution  of
1000 mg/L of 1,4-dichloronaphthalene and dilute  it to 100 ng//iL.  Use a spiking
volume of  100 ^l for  a  1 L  aqueous  sample.    If matrix  interferences  are  a
                                   8121 - 5                         Revision 0
                                                                September 1994

-------
problem,  two  alternative  surrogates  are:  alpha,  2,6-trichlorotoluene  or
2,3,4,5,6-pentachlorotoluene.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material to this Chapter, Organic Analytes, Sec.
4.1.

      6.2   Extracts must  be stored  at  4 °C and  analyzed  within 40  days  of
extraction.


7.0   PROCEDURE

      7.1   Extraction and Cleanup:

            7.1.1 Refer to Chapter Two and Method 3500 for guidance on choosing
      the  appropriate  extraction  procedure.   In  general,  water  samples  are
      extracted at a  neutral, or as is, pH with methylene chloride, using either
      Method 3510  or  3520.   Solid samples are extracted using  either Methods
      3540,  3541,  or  3550  with  methylene  chloride/acetone  (1:1)  as  the
      extraction solvent.

            7.1.2 If required,  the samples may be cleaned up using Method 3620
      (Florisil)  and/or Method  3640  (Gel  Permeation  Chromatography).    See
      Chapter Two, Sec.  2.3.2 and  Method  3600 for  general  guidance on cleanup
      and method selection.  Method 3660 is  used for sulfur removal.

            7.1.3 Prior to gas chromatographic analysis,  the extraction solvent
      must exchanged  into  hexane using the Kuderna-Danish  concentration step
      found in any of the extraction  methods.  Any methylene chloride remaining
      in the extract will cause a very broad solvent peak.

      7.2   Gas Chromatographic Conditions:

            7.2.1 Retention  time  information  for  each  of the  analytes  is
      presented in Tables 3 and  4.  The  recommended GC operating conditions are
      provided  in  Tables  5  and 6.   Figures 1,  2  and 3  illustrate  typical
      chromatography of the method  analytes  for both the  single column approach
      and the dual column approach  when  operated at the  conditions specified in
      Tables 5 and 6.

      7.3   Calibration:

            7.3.1 Prepare  calibration  standards  using the  procedures  in Sec.
      5.0.   Refer  to  Method   8000  for  proper calibration  procedures.  The
      procedure for internal or external calibration may be used.

            7.3.2 Refer to Method 8000 for the establishment of retention time
      windows.
                                   8121 - 6                         Revision 0
                                                                September 1994

-------
7.4   Gas chromatographic analysis:

      7.4.1 Method 8000  provides  instructions  on  the analysis sequence,
appropriate  dilutions,  establishing daily  retention time  windows,  and
identification criteria.

      7.4.2 Automatic injections of 1 pi are recommended.  Hand injections
of no more than 2 /zL may be  used if the analyst demonstrates quantitation
precision of < 10 percent relative standard  deviation.  The  solvent flush
technique may be used if the amount of solvent is kept at a minimum.  If
the  internal  standard calibration technique is  used,  add  10 jitL of the
internal standard to each ml of sample extract prior to injection.

      7.4.3 Tentative identification of an analyte occurs when  a peak from
a sample extract falls within the daily retention time window.

      7.4.4 Validation   of   gas   chromatographic   system   qualitative
performance: Use  the  midconcentration  standards  interspersed throughout
the  analysis sequence  (Sec.  7.3)  to evaluate this criterion.   If any of
the  standards fall outside their daily retention time windows, the system
is out of control. Determine the cause of the problem and correct it (see
Sec. 7.5).

      7.4.5 Record the  volume  injected to  the nearest  0.05 /uL  and the
resulting  peak  size  in  peak height or  area  units.   Using  either the
internal or the  external  calibration procedure (Method  8000), determine
the  identity and  the quantity  of each  component  peak  in  the sample
chromatogram which  corresponds  to the compounds used  for calibration
purposes.  See Method 8000 for calculation equations.

      7.4.6 If the responses exceed the linear range of the system, dilute
the  extract and reanalyze. Peak height measurements are recommended over
peak  area  integration   when  overlapping  peaks   cause   errors   in  area
integration.

      7.4.7 If partially overlapping or coeluting  peaks are  found, change
columns  or  try  a  GC/MS technique (see  Sec.  8.7 and  Method  8270).
Interferences that prevent analyte identification  and/or quantitation may
be removed by the cleanup techniques mentioned above.

      7.4.8 If the peak response is less than 2.5  times the baseline noise
level, the validity of the quantitative result may be questionable.  The
analyst should consult with  the  source of the sample to determine whether
further concentration of the sample is  warranted.

7.5   Instrument Maintenance:

      7.5.1 Injection of sample extracts  from waste sites often leaves a
high boiling residue  in:  the injection port  area,  splitters when used, and
the  injection  port  end  of  the chromatographic   column.    This  residue
effects chromatography in many  ways  (i.e.,  peak  tailing,  retention time
shifts, analyte degradation,  etc.)  and, therefore, instrument maintenance
                             8121 - 7                         Revision 0
                                                          September 1994

-------
      is very important.  Residue buildup in a splitter may limit flow through
      one leg and therefore change the split ratios.  If this occurs during an
      analytical run, the quantitative data may  be  incorrect.   Proper cleanup
      techniques will minimize the problem and  instrument QC will indicate when
      instrument maintenance is required.

            7.5.2 Suggested chromatograph maintenance: Corrective measures may
      require any one or more  of the  following  remedial actions.  Also see Sec.
      7  in  Method  8000  for   additional  guidance  on corrective  action  for
      capillary columns and the injection port.

                  7.5.2.1     Splitter connections:  For dual columns which are
            connected using a  press-fit  Y-shaped  glass splitter or a Y-shaped
            fused-silica connector, clean and deactivate the splitter or replace
            with a cleaned  and  deactivated  splitter.   Break off the first few
            inches (up to one  foot)  of the  injection  port  side of the column.
            Remove  the  columns   and  solvent  backflush   according  to  the
            manufacturer's instructions.  If these procedures fail to eliminate
            the degradation problem,  it may be necessary to  deactivate the metal
            injector body and/or replace the columns.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and Method 8000 for  specific quality control
procedures.   Quality control to validate  sample extraction  is covered in Method
3500 and in the extraction method utilized.   If extract cleanup was performed,
follow the QC in Method 3600 and in the specific cleanup method.

      8.2   Quality control  required  to evaluate the GC system operation is found
in Method 8000, Sec. 8.3.

      8.3   Calculate surrogate standard  recoveries for all  samples, blanks, and
spikes.   Determine if  the  recovery   is  within limits (limits  established by
performing QC procedures outlined in Method 8000, Sec. 8).   If the recovery is
not within limits, the following are required:

            8.3.1 Check  to be  sure there  are  no  errors in  calculations,
      surrogate  solutions   and  internal   standards.    Also, check  instrument
      performance.

            8.3.2 Recalculate  the data and/or  reanalyze  the extract if any of
      the above checks reveal  a problem.

            8.3.3 Reextract and  reanalyze the  sample  if  none of the above are
      a problem, or  flag the data as  "estimated concentrations".

      8.4   Data from systems  that  automatically identify target analytes  on the
basis  of  retention time or retention time indices  should be  reviewed  by an
experienced analyst before  they are  reported.

      8.5   When using the internal standard calibration technique, an internal
standard peak area check must be performed on all samples.  The internal standard

                                   8121  - 8                         Revision 0
                                                                September 1994

-------
must be evaluated for acceptance by determining whether the measured area for the
internal standard deviates by more  than 50 percent from the average area for the
internal standard in the calibration standards.  When the  internal  standard peak
area is outside that limit, all samples that fall outside the QC criteria must
be reanalyzed.

      8.6   Include a mid-concentration calibration standard after each group of
20  samples   in   the   analysis  sequence.     The  response  factors  for  the
mid-concentration calibration must be  within +  15 percent of the average values
for the multiconcentration calibration.   When the response factors  fall outside
that  limit,   all  samples  analyzed after  that  mid-concentration calibration
standard must be reanalyzed after performing instrument maintenance to correct
the usual source  of  the problem.   If  this  fails to correct the problem, a new
calibration curve must be established.

      8.7   GC/MS confirmation:

            8.7.1 GC/MS techniques  should  be judiciously  employed to support
      qualitative  identifications  made  with this method.   Follow  the GC/MS
      operating  requirements  specified  in  Method 8270.   Ensure  that there is
      sufficient concentration of the  analyte(s) to be confirmed,  in the extract
      for GC/MS analysis.

            8.7.2 When  available,   chemical  ionization  mass  spectra  may  be
      employed to aid  in the qualitative identification process.

            8.7.3 To confirm  an  identification of a  compound,  the background
      corrected mass spectrum  of the compound must be obtained from the sample
      extract  and must  be  compared  with  a mass  spectrum from a  stock  or
      calibration standard analyzed under the same chromatographic conditions.
      At  least  25  ng  of material  should  be  injected  into  the  GC/MS.   The
      identification  criteria  specified  in  Method   8270 must   be  met  for
      qualitative confirmation.

                  8.7.3.1      Should  the   MS   procedure  fail   to  provide
            satisfactory  results,  additional   steps  may  be  taken  before
            reanalysis.  These steps may  include the use  of alternate packed or
            capillary GC columns or additional  sample cleanup.


9.0   METHOD PERFORMANCE

      9.1   The MDL is defined in Chapter One.   The MDLs  listed in  Table 1 were
obtained by using organic-free reagent water.  Details  on  how to determine MDLs
are given in Chapter One.  The MDLs actually achieved in a given analysis will
vary since they depend on instrument sensitivity and matrix effects.

      9.2   This  method   has   been  tested   in  a  single  laboratory   by  using
organic-free reagent water,  sandy  loam samples  and  extracts  which were spiked
with the  test  compounds  at one concentration.   Single-operator  precision  and
method accuracy were  found to be related to the concentration of compound and the
type of matrix.


                                   8121  - 9                         Revision 0
                                                                September 1994

-------
      9.3   Single  laboratory  accuracy  data  were  obtained  for  chlorinated
hydrocarbons in a clay  soil.  The spiking concentrations ranged from 500 to 5000
jug/kg,  depending on the  sensitivity of the  analyte to the electron capture
detector.  The spiking  solution was mixed into the soil  during addition and then
immediatly transferred to the extraction device and immersed in the extraction
solvent.   The  spiked  sample  was then  extracted by  Method 3541  (Automated
Soxhlet).  The data represents a single determination.  Analysis was by capillary
column gas chromatography/electron capture detector following Method 8121 for the
chlorinated hydrocarbons.  These data are listed  in Table 9 and were taken from
Reference 4.
10.0  REFERENCES

1.    Lopez-Avila,  V.,  N.S.  Dodhiwala,  and  J.  Milanes,   "Single  Laboratory
      Evaluation of Method 8120, Chlorinated Hydrocarbons", 1988, EPA Contract
      Numbers 68-03-3226 and 68-03-3511.

2.    Glazer, J.A., G.D. Foerst, G.D.  McKee, S.A.  Quave, and W.L. Budde, "Trace
      Analyses for Wastewaters," Environ.  Sci. and Technol.  15:1426-1431, 1981.

3.    Lopez-Avila, V.;  Baldin,  E.;  Benedicto,  J; Milanes,  J.;  Beckert,  W. F.
      "Application of Open-Tubular Columns to SW 846 GC Methods"; final report
      to  the U.S.  Environmental  Protection  Agency  on Contract  68-03-3511;
      Mid-Pacific Environmental Laboratory, Mountain View,  CA, 1990.

4.    Lopez-Avila, V. (Beckert, W., Project Officer), "Development of a Soxtec
      Extraction  Procedure for  Extracting Organic  Compounds from  Soils  and
      Sediments",  EPA  600/X-91/140,  US EPA, Environmental  Monitoring Systems
      Laboratory-Las Vegas, October 1991.
                                   8121 - 10                         Revision 0
                                                                September  1994

-------
                             TABLE 1
      METHOD DETECTION LIMITS FOR CHLORINATED HYDROCARBONS
                SINGLE COLUMN METHOD OF ANALYSIS
Compound name
Benzal chloride
Benzotri chloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachl orobutadi ene
a~Hexachlorocyclohexane (a-BHC)
/3-Hexachlorocyclohexane ((8-BHC)
-y-Hexachlorocyclohexane (7-BHC)
5 -Hexachl orocyclohexane (
-------
                                 TABLE 2
     ESTIMATED  QUANTITATION  LIMIT  (EQL)  FACTORS FOR VARIOUS MATRICES'
      Matrix                                                Factor
Ground water                                                     10
Low-concentration soil by ultrasonic extraction                 670
  with GPC cleanup
High-concentration soil and sludges by ultrasonic            10,000
  extraction
Waste not nriscible with water                               100,000
a   EQL  = [Method detection limit (see Table 1)] x [Factor found  in  this
    table].   For nonaqueous  samples,  the factor is on  a  wet-weight basis.
    Sample EQLs are highly matrix-dependent.  The EQLs  listed  herein are
    provided for guidance and  may not always be achievable.
                                   8121  - 12                        Revision 0
                                                                September 1994

-------
                                 TABLE 3
 GAS  CHROMATOGRAPHIC  RETENTION  TIMES  FOR  CHLORINATED  HYDROCARBONS:  SINGLE
                        COLUMN METHOD OF ANALYSIS
Compound name
Retention time (min)
DB-210"DB-WAXb
Benzal chloride
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Hexachlorobenzene
Hexachlorobutadiene
a-BHC
K-BHC
6-mt
Hexachl orocycl opentadi ene
Hexachloroethane
Pentachlorobenzene
1,2, 3 ,4-Tetrachl orobenzene
1,2,4,5-Tetrachlorobenzene
1,2,3 , 5-Tetrachl orobenzene
1 , 2 , 4-Tr i chl orobenzene
1,2, 3 -Trichl orobenzene
1, 3, 5-Tri chl orobenzene
6.86
7.85
4.59
13.45
4.44
3.66
3.80
19.23
5.77
25.54
24.07
26.16
8.86
3.35
14.86
11.90
10.18
10.18
6.86
8.14
5.45
15.91
15.44
10.37
23.75
9.58
7.73
8.49
29.16
9.98
33.84
54.30
33.79
c
8.13
23.75
21.17
17.81
17.50
13.74
16.00
10.37
Internal Standards

2,5-Dibromotoluene
1,3,5-Tri bromobenzene
a,a'-Di bromo-meta-xylene

Surrogates
 9.55
11.68
18.43
or,2,6-Trichlorotoluene        12.96
1,4-Dichloronaphthalene       17.43
2,3,4,5,6-Pentachlorotoluene  18.96
18.55
22.60
35.94
                  22.53
                  26.83
                  27.91
   GC  operating conditions:    30  m  x 0.53  mm  ID  DB-210  fused-silica
   capillary column; 1 /xm film thickness;  carrier gas helium at 10 mL/min;
   makeup gas is nitrogen  at  40 mL/min; temperature  program  from 65°C to
   175°C (hold 20 minutes) at 4°C/min;  injector temperature 220°C; detector
   temperature 250°C.

   GC  operating conditions:    30  m  x 0.53  mm  ID  DB-WAX  fused-silica
   capillary column; 1 /urn film thickness;  carrier gas helium at 10 mL/min;
   makeup gas is nitrogen  at  40 mL/min; temperature  program  from 60°C to
   170°C (hold 30 minutes) at 4°C/min;  injector temperature 200°C; detector
   temperature 230°C.

   Compound decomposes on-column.
                                   8121  -  13
                                      Revision  0
                                  September 1994

-------
                                TABLE 4
            RETENTION TIMES OF THE CHLORINATED HYDROCARBONS'
                     DUAL COLUMN METHOD OF ANALYSIS
Compound
1,3-Dichlorobenzene
1,4-Dichlorobenzene
Benzyl chloride
1,2-Dichlorobenzene
Hexachloroethane
1,3, 5-Trichl orobenzene
Benzal chloride
1, 2, 4-Trichl orobenzene
1,2, 3 -Tri chl orobenzene
Hexachlorobutadiene
Benzotrichloride
1,2,3 , 5-Tetrachl orobenzene
1,2,4 , 5-Tetrachl orobenzene
Hexachl orocycl opentadi ene
1,2,3 , 4-Tetrachl orobenzene
2-Chl oronaphthal ene
Pentachl orobenzene
a-BHC
Hexachl orobenzene
0-BHC
7-BHC
5-BHC
DB-5
RT(min)
5.82
6.00
6.00
6.64
7.91
10.07
10.27
11.97
13.58
13.88
14.09
19.35
19.35
19.85
21.97
21.77
29.02
34.64
34.98
35.99
36.25
37.39
DB-1701
RT(min)
7.22
7.53
8.47
8.58
8.58
11.55
14.41
14.54
16.93
14.41
17.12
21.85
22.07
21.17
25.71
26.60
31.05
38.79
36.52
43.77
40.59
44.62
Internal Standard
1,3,5-Tribromobenzene                         11.83          13.34

Surrogate
1,4-Dichloronaphthalene                       15.42          17.71

"The  GC operating conditions  were  as follows:  30 m x  0.53  mm  ID  DB-5
(0.83-|um   film thickness) and  30  m x 0.53  mm  ID  DB-1701  (1.0 p.m  film
thickness)  connected to  an 8-in injection tee (Supelco Inc.).  Temperature
program: 80°C (1.5 min hold) to 125°C (1 min hold) at 2'C/min then to 240eC
(2 min hold) at 5°C/min;  injector temperature 250°C; detector  temperature
320°C; helium  carrier gas 6 mL/min; nitrogen makeup gas 20 mL/min.
                                   8121  -  14                        Revision 0
                                                                September 1994

-------
                              TABLE 5
             GC OPERATING CONDITIONS FOR CHLOROHYDROCARBONS
                    SINGLE COLUMN METHOD OF ANALYSIS
Column  1:  DB-210    30 m  x  0.53  mm  ID  fused-silica  capillary  column
chemically bonded with trifluoropropyl methyl silicone
    Carrier gas  (He)     10 mL/min
    Column  temperature:
                  Initial temperature
                  Temperature program
                  Final temperature
    Injector temperature      220°C
    Detector temperature      250"C
    Injection volume           1-2 jtiL
65'C
658C to 175°C at 4°C/min
175°C, hold 20 minutes.
    Column  2:   DB-WAX    30  m x 0.53 mm ID  fused-silica  capillary  column
    chemically bonded with polyethylene glycol
    Carrier gas  (He)     10 mL/min
    Column  temperature:
                  Initial temperature
                  Temperature program
                  Final  temperature
    Injector temperature      200°C
    Detector temperature      230°C
    Injection volume           1-2 /iL
60°C
60°C to 170°C at 4°C/min
170°C, hold 30 minutes.
                                  8121  -  15
                        Revision  0
                    September  1994

-------
                              TABLE 6
           GC OPERATING CONDITIONS FOR CHLORINATED HYDROCARBONS
                     DUAL COLUMN METHOD OF ANALYSIS
Column 1:
Column 2:
                  Type:  DB-1701 (J&W Scientific) or equivalent
                  Dimensions:  30 m x 0.53 mm ID
                  Film Thickness: 1.0 (jum)
                  Type:  DB-5 (J&W Scientific) or equivalent
                  Dimensions:  30 m x 0.53 mm ID
                  Film Thickness: 0.83
Carrier gas flowrate (mL/min):  6 (Helium)

Makeup gas flowrate (mL/min):  20 (Nitrogen)

Temperature program:  80° C (1.5 mi n  hold)  to 125°C (1 min hold) at 2°C/min
then to 240°C (2 min hold) at 5°C/min.

Injector temperature:  250°C

Detector temperature:  320°C

Injection volume:  2 juL

Solvent:  Hexane

Type of injector:  Flash vaporization

Detector type:  Dual ECD

Range:  10

Attenuation:  32 (DB-1701J/32  (DB-5)

Type of splitter:  Supelco 8-in injection tee
                                   8121 - 16                         Revision 0
                                                                 September  1994

-------
                                 TABLE 7

         SUGGESTED CONCENTRATIONS FOR THE CALIBRATION SOLUTIONS8


                              Concentration (ng//nL)
Benzal chloride
Benzotrichloride
Benzyl chloride
2-Chloronaphthalene
1,2-Dichlorobenzene
1 , 3 -Di chl orobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachlorobutadiene
a-BHC
/3-BHC
T-BHC
6-BHC
Hexachl orocycl opentad i ene
Hexachl oroethane
Pentachl orobenzene
1,2,3 , 4-Tetrachl orobenzene
1,2, 4, 5-Tetrachl orobenzene
1 ,2,3,5-Tetrachlorobenzene
1, 2, 4-Tri chl orobenzene
1 , 2 , 3-Tri chl orobenzene
1, 3, 5-Tri chl orobenzene
0.1
0.1
0.1
2.0
1.0
1.0
1.0
0.01
0.01
0.1
0.1
0.1
0.1
0.01
0.01
0.01
0.1
0.1
0.1
0.1
0.1
0.1
0.2
0.2
0.2
4.0
2.0
2.0
2.0
0.02
0.02
0.2
0.2
0.2
0.2
0.02
0.02
0.02
0.2
0.2
0.2
0.2
0.2
0.2
0.5
0.5
0.5
10
5.0
5.0
5.0
0.05
0.05
0.5
0.5
0.5
0.5
0.05
0.05
0.05
0.5
0.5
0.5
0.5
0.5
0.5
0.8
0.8
0.8
16
8.0
8.0
8.0
0.08
0.08
0.8
0.8
0.8
0.8
0.08
0.08
0.08
0.8
0.8
0.8
0.8
0.8
0.8
1.0
1.0
1.0
20
10
10
10
0.1
0.1
1.0
1.0
1.0
1.0
0.1
0.1
0.1
1.0
1.0
1.0
1.0
1.0
1.0
Surrogates

a,2,6-Trichlorotoluene        0.02   0.05    0.1      0.15    0.2
1,4-Dichloronaphthalene       0.2    0.5     1.0      1.5     2.0
2,3,4,5,6-Pentachlorotoluene  0.02   0.05    0.1      0.15    0.2
    One  or  more  internal  standards  should  be  spiked prior  to  GC/ECD
    analysis  into all  calibration solutions.   The spike concentration  of
    the  internal  standards  should  be kept constant  for all calibration
    solutions.
                                  8121  -  17
    Revision 0
September 1994

-------
                                 TABLE  8
              ELUTION PATTERNS OF CHLORINATED HYDROCARBONS
  FROM THE FLORISIL COLUMN BY ELUTION WITH PETROLEUM ETHER (FRACTION 1)
           AND 1:1 PETROLEUM ETHER/DIETHYL ETHER (FRACTION 2)
Compound
Benzal chlorided
Benzotri chloride
Benzyl chloride
2-Chloronaphthalene
1 ,2-Dichlorobenzene
1 , 3-Di chl orobenzene
1,4-Dichlorobenzene
Hexachl orobenzene
Hexachl orobutadi ene
a-BHC
/3-BHC
y-BHC
S-BHC
Hexachl orocycl opentadi ene
Hexachl oroethane
Pentachl orobenzene
1,2,3 , 4-Tetrachl orobenzene
1,2,4 , 5-Tetrachl orobenzene"
1,2,3 , 5-Tetrachl orobenzene6
1 , 2 , 4-Tri chl orobenzene
1,2, 3 -Tri chl orobenzene
1 , 3 , 5-Tri chl orobenzene
Amount
(M9)
10
10
100
200
100
100
100
1.0
1.0
10
10
10
10
1.0
1.0
1.0
10
10
10
10
10
10
Recovery
Fraction 1"
0
0
82
115
102
103
104
116
101




93
100
129
104
102
102
59
96
102
(percent)"
Fraction 2°
0
0
16






95
108
105
71









a   Values  given  represent  average  values  of duplicate  experiments.
b   Fraction  1  was  eluted with  200  mL  petroleum ether.
c   Fraction  2  was  eluted with 200 mL petroleum ether/diethyl  ether (1:1).
d   This    compound  coelutes   with   1,2,4-trichlorobenzene;    separate
    experiments were performed with benzal  chloride to verify that  this
    compound  is not recovered from the  Florisil  cleanup in either fraction.
e   This  pair  cannot  be resolved  on  the  DB-210 fused-silica capillary
    columns.
                                   8121  -  18
    Revision 0
September 1994

-------
                                    TABLE  9
             SINGLE  LABORATORY  ACCURACY  DATA  FOR THE  EXTRACTION  OF
         CHLORINATED HYDROCARBONS  FROM SPIKED CLAY SOIL  BY METHOD 3541
                             (AUTOMATED SOXHLET)"
Compound Name                 Spike Level                 %  Recovery
                                 jug/kg                DB-5        DB-1701
1,3-Dichlorobenzene              5000                   b            39
1,2-Dichlorobenzene              5000                  94            77
Benzal chloride                   500                  61            66
Benzotrichloride                  500                  48            53
Hexachlorocyclopentadiene         500                  30            32
Pentachlorobenzene                500                  76            73
alpha-BHC                         500                  89            94
delta-BHC                         500                  86             b
Hexachlorobenzene                 500                  84            88
a     The  operating  conditions  for  the  automated Soxhlet  were  as  follows:
      immersion time 45 min; extraction time 45 min;  the  sample size  was 10 g
      clay soil,  extraction solvent,  1:1 acetone/hexane.  No equilibration time
      following spiking.

b     Not able to determine because of interference.

Data taken from Reference 4.
                                  8121  -  19                         Revision 0
                                                                September 1994

-------
1
3 ;
•
n

N J

I
21 11

14
Iff


                                                     10  ia 1113
                                                      ll
10
                                       If

                                   TIMi(mln)
ao
                            as
30
Figure 1.   GC/ECD chromatogram of Method 8121 composite standard analyzed on a
            30 m  x  0.53  mm  ID DB-210  fused-silica  capillary column.    GC
            operating conditions are given  in Section 7.4.   See Table  3  for
            compound Identification.
                                  8121 - 20
                                      Revision 0
                                  September 1994

-------
                                 4
                                 16
                                                        10
                                              ft
                                              13
                                              A
             10
15
20
25     30

TIME (min)
35
                                                     40
                                        45
                                        50
55
Figure 2.   6C/ECD chromatogram of Method 8121 composite standard analyzed on a
            30 ro  x  0.53 mm  ID  DB-WAX fused-silica  capillary  column.    GC
            operating conditions  are  given  in Section 7.4.   See Table  3  for
            compound identification.
                                   8121  -  21
                                                 Revision  0
                                             September 1994

-------
                                                             11  If
                                      OB-1701
           i
                         to
                         •
                         T    .11    14 It     tl  IS     It  tt     tt   II  fl    20
u
                  JU
Figure 3.   GC/ECD chromatogram  of chlorinated  hydrocarbons analyzed on  a DB
            5/DB 1701 fused-silica, open-tubular column pair.  The GC operating
            conditions were as follows:   30 m x 0.53 mm ID  DB 5  (0.83 /*m film
            thickness) and 30  m  x 0.53 mm  ID  DB 1701 (1.0  urn film thickness)
            connected  to  an  8 In Injection tee  (Supelco  Inc.).   Temperature
            program: 80°C (1.5 m1n hold) to 125°C (1 m1n hold) at 2°C/min,  then
            to 240°C (2 m1n hold) at 5°C/m1n.
                                   8121 - 22
                        Revision 0
                    September 1994

-------
                                               METHOD 8121

CHLORINATED  HYDROCARBONS BY  GAS  CHROMATOGRAPHY:  CAPILLARY COLUMN  TECHNIQUE
                                               7.1.1
                                                      I
                                             7.12 Add appropriate spiking
                                             coffipoundi to sample prior
                                               to extraction procedure
                                              7.2 Exchange oxtracbon
                                              solvent to hexane during
                                                 K-D procedure
                                            7.2.1 Following concentration of
                                             mMhylww chkxid* allow K-0
                                             •ppiratu* to drain and cool
                                          122. InctMW tMnptratun of hot
                                           nwtar btlh; *dd hMww; tttrndi
                                       Snydv cohjtnn; plae* •opwikji on «MMr
                                                      §^ rpfiwa from
                                                 vMttr b«th; cool
                                           72.3 Remov* column; rint* leak
                                            •nd jointt iMlth hcxin*; adjiMt
                                                  extract volume
         7 JS Choo*» «ppfopn«t« cleanup
            technique, if neceteary;
        HUORMH cleanup ie recommended.
                                      723 Will further
                                       proceaiingbe
                                      performed iMtNn
                                        MM days?
 72.3 Tranefer extract to
Teflon sealed screw^ap
           Refer to Method 3620 or to
                Section 7.32
             7.3.3 GPC
             cleanup
             required?
                        7.3.4 Elemental
                         sulfur removal
                           required?
                                     7.3.4 Refer to

                                      Section 7.3  '
7.33 Refer to
Method 3640
                                                8121  -  23
                                                                                       Revision  0
                                                                                 September  1994

-------
   METHOD  8121
   (continued)
         O
1
1
7.2.3 Stopper concentrator
and refrigerate
            I
7.4.1 Set column 1 conditions
 7.4.2 Set column 2 conditions
  7.5.1 Refer to Method 8000 for
  caiibration techniques; select
 lowest point on calibration curve
  7.5.2 Choose and perform
 internal or external calibration
    (refer to Method 8000)
  7.6.1 Add internal standard
         if necessary
7.6.2 Establish daily retention time
  windows, analysis sequence,
dilutions, and identification criteria
         O
         8121  -  24
      Revision  0
September  1994

-------
                        METHOD 8121
                        (concluded)
          0
  7.6.3 Record sample volume
  injected and resulting peak
           sizes
             i
  7.6.4 Determine identity and
quantity of each component peak
 that corresponds to compound
      used for calibration
           7.6.5
         Does peak
       exceed working
          range of
          system?
7.6.5 Dilute extract; reanalyze
 7.6.6 Compare standard and
   sample retention times;
     identify compounds
                            8121  -  25
                           Revision  0
                     September  1994

-------
                                 METHOD 8140

                         ORGANOPHOSPHORUS PESTICIDES
1.0  SCOPE AND APPLICATION

     1.1  Method 8140 is a gas  chromatographic  (GC)  method used to determine
the concentration of various organosphosphorus  pesticides.   Table 1 indicates
compounds that may be determined by this method and lists the method detection
limit for each  compound  in  reagent  water.    Table  2  lists the practical
quantisation limit (PQL) for other matrices.

     1.2  When Method 8140  is  used  to  analyze unfamiliar samples, compound
identifications should be  supported  by  at  least two additional qualitative
techniques if mass spectroscopy  is  not  employed.   Section 8.4 provides gas
chromatograph/mass  spectrometer   (GC/MS)   criteria   appropriate   for  the
qualitative confirmation of compound identifications.


2.0  SUMMARY OF METHOD

     2.1  Method  8140  provides   gas   chromatographic  conditions  for  the
detection of ppb  levels  of  organophosphorus  pesticides.  Prior to analysis,
appropriate sample extraction techniques must be  used.  Both neat and diluted
organic  liquids   (Method  3580,  Waste  Dilution)  may  be  analyzed by direct
injection.  A 2-  to  5-uL  aliquot  of  the  extract  is  injected into a gas
chromatograph, and compounds in  the  GC  effluent  are  detected with a flame
photometric or thermionic detector.

     2.2  If interferences are encountered   in  the  analysis, Method 8140 may
also be  performed on extracts  that  have   undergone cleanup using Method 3620
and/or Method 3660.


3.0  INTERFERENCES

     3.1  Refer to Methods 3500  (Section 3.5, in particular), 3600, and 8000.

     3.2  The use of Florlsil cleanup materials  (Method 3620) for some of the
compounds in this method has  been  demonstrated to yield recoveries less than
85% and  is therefore not recommended for  all  compounds.  Refer to Table 2 of
Method 3620 for recoveries  of  organophosphorous  pesticides as a function of
Florisil  fractions.    Use  of  phosphorus-  or  halogen-specific  detectors,
however, often obviates the necessity   for  cleanup for relatively clean sample
matrices.   If  particular  circumstances   demand  the  use  of an alternative
cleanup  procedure,  the  analyst  must  determine  the  elution  profile  and
demonstrate that  the recovery of each analyte is no less than 85%.
                                   8140 -  1
                                                         Revision      0
                                                         Date  September 1986

-------
TABLE 1.  GAS CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION  LIMITS  FOR
          ORGANOPHOSPHOROUS PESTICIDES3
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton-0
Demeton-S
Diazinon
Dichlorvos
Disulfoton
Ethoprop
Fensulfothion
Fenthion
Merphos
Mevinphos
Naled
Parathion methyl
Phorate
Ronnel
Stirophos (Tetrachlorvinphos)
Tokuthion (Prothiofos)
Trichloronate
GC
col umn"3
la
la
2
la
la
la
2
Ib, 3
la
2
la
la
2
Ib
3
2
la
2
Ib, 3
la
la
Retention
time
(min)
6.80
4.23
6.16
11.6
2.53
1.16
7.73
0.8, 1.50
2.10
3.02
6.41
3.12
7.45
2.41
3.28
3.37
1.43
5.57
8.52, 5.51
3.40
2.94
Method
detection
limit (ug/L)
1.5
0.15
0.3
1.5
0.25
0.25
0.6
0.1
0.20
0.25
1.5
0.10
0.25
0.3
0.1
0.03
0.15
0.3
5.0
0.5
0.15
      Development of  Analytical   Test   Procedures   for  Organic  Pollutants in
      Wastewater;  Report for EPA Contract 68-03-2711  (in preparation).

      bSee Sections 4.2.1 and 7.2 for column  descriptions and conditions.
                                   8140 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
TABLE 2.  DETERMINATION OF PRACTICAL QUANTITATION LIMITS (PQL) FOR VARIOUS
          MATRICES3


    Matrix                                                   Factorb
Ground water                                                     10
Low-level soil by sonication with GPC cleanup                   670
High-level soil and sludges by sonication                    10,000
Non-water miscible waste                                    100,000


     aSample PQLs are highly  matrix-dependent.    The  PQLs listed herein are
     provided for guidance and may not always be achievable.

     bPQL = [Method detection limit (Table 1)] X [Factor (Table 2)].  For non-
     aqueous samples, the factor is on a wet-weight basis.
                                  8140 - 3
                                                         Revision
                                                         Date  September 1986

-------
     3.3  Use of a  flame  photometric  detector  1n  the phosphorus mode will
minimize  interferences  from  materials   that  do  not  contain  phosphorus.
Elemental sulfur, however,  may  interfere  with  the determination of certain
organophosphorus pesticides by flame  photometric  gas chromatography.  Sulfur
cleanup using Method 3660 may alleviate this interference.

     3.4  A  halogen-specific  detector  (i.e.,  electrolytic  conductivity or
microcoulometric) is very selective  for the halogen-containing pesticides and
is recommended for use with dichlorvos, naled, and stirophos.


4.0  APPARATUS AND MATERIALS

     4.1  Gas   chrpmatograph;      Analytical   system   complete   with  gas
chromatograph suitable for on-column  Injections and all required accessories,
including detectors, column supplies, recorder,  gases,  and syringes.  A data
system for measuring peak areas and/or peak heights is recommended.

          4.1.1  Columns:

               4.1.1.1  Column la and Ib:    1.8-m  x  2-mm I.D. glass, packed
          with 5% SP-2401 on Supelcoport, 100/120 mesh  (or equivalent).

               4.1.1.2  Column 2:  1.8-m x 2-mm I.D. glass, packed with 3% SP-
          2401 on Supelcoport, 100/120 mesh (or equivalent).

               4.1.1.3  Column 3:  50-cm x l/8-1n O.D.  Teflon, packed with 15%
          SE-54 on Gas Chrom Q, 100/120 mesh  (or equivalent).

          4.1.2  Detectors:  The following  detectors  have proven effective in
     analysis for the  analytes listed  1n Table 1 and  were used  to develop the
     accuracy and precision statements  in Section 9.0.

               4.1.2.1   Phosphorus-specific:      Nitrogen/Phosphorus    (N/P),
          operated  in  phosphorus-sensitive mode.

               4.1.2.2  Flame  Photometric  (FPD):     FPD  is  more selective for
          phosphorus than the  N/P.

               4.1.2.3  Halogen-specific:       Electrolytic    conductivity  or
          microcoulometric.   These   are  very  selective  for those  pesticides
          containing halogen  substituents.

     4.2 Balance;   analytical,  capable of  accurately weighing  to  the nearest
 0.0001 g.

     4.3 Vials;   Amber glass,  10-   to 15-mL  capacity with Teflon-lined screw-
 cap.

     4.4  Kuderna-Danish (K-D)  apparatus;

          4.4.1   Concentrator tube:   10-mL, graduated (Kontes K-570050-1025 or
     equivalent).   Ground-glass   stopper  is   used   to  prevent  evaporation of
     extracts

                                   8140 - 4
                                                          Revision      0
                                                          Date  September 1986

-------
          4.4.2  Evaporation   flask:       500-mL   (Kontes    K-570001-500   or
     equivalent).  Attach to concentrator tube with  springs.

          4.4.3  Snyder column:    Three-ball   macro  (Kontes  K-503000-0121  or
     equivalent).

          4.4.4  Snyder  column:    Two-ball   micro   (Kontes   K-569001-0219  or
     equivalent).

     4.5  Boiling chips;  Solvent extracted,  approximately 10/40 mesh (silicon
carbide or equivalent).

     4.6  Water  bath;    Heated,  with  concentric   ring  cover,  capable   of
temperature control (+5*C).  The bath should  be used in a hood.

     4.7  Microsyringe;  10-uL.

     4.8  Syringe;  5-mL.

     4.9  Volumetric flasks;  10-, 50-, and 100-mL,  ground-glass stopper.


5.0  REAGENTS

     5.1  Solvents;    Hexane,   acetone,  isooctane  (2,2,4-trimethylpentane)
(pesticide quality or equivalent).

     5.2  Stock  standard solutions;

          5.2.1  Prepare stock standard solutions by accurately weighing about
     0.0100 g of pure  material.    Dissolve  the  material in hexane or other
     suitable solvent  and  dilute  to  volume  in  a  10-mL volumetric flask.
     Larger volumes can  be  used  at  the  convenience  of  the  analyst.   If
     compound purity is certified at  96%  or  greater, the weight can be used
     without correction to calculate the  concentration of the stock standard.
     Commercially prepared stock standards can be used at any concentration if
     they are certified by the manufacturer or by an independent source.

          5.2.2  Transfer  the   stock  standard  solutions  into Teflon-sealed
     screw-cap bottles.  Store  at 4*C  and protect  from light.   Stock standard
     solutions should  be  checked  frequently  for  signs  of   degradation or
     evaporation,  especially  just  prior  to  preparing calibration standards
     from them.

          5.2.3  Stock standard  solutions must be   replaced after one year, or
     sooner if comparison  with  check standards indicates a problem.

     5.3  Calibration  standards;  Calibration  standards  at a minimum of five
concentration  levels  for each  parameter  of interest should be prepared through
dilution of the  stock  standards  with   isooctane.    One of the concentration
levels  should be at   a  concentration  near,  but   above, the method detection
                                  8140 - 5
                                                         Revision      0
                                                         Date  September 1986

-------
limit.  The remaining concentration  levels  should correspond to the expected
range of concentrations found  in  real   samples  or should define the working
range of the GC.  Calibration standards   must be replaced after six months,  or
sooner if comparison with check standards indicates a problem.

     5.4  Internal standards (If Internal  standard  calibration is used);  To
use this approach, the analyst must select one or more internal standards that
are similar in analytical behavior to  the compounds of interest.  The analyst
must further demonstrate that the measurement  of the internal standard is not
affected by method or matrix interferences.   Because of these limitations,  no
internal standard can be suggested that is applicable to all samples.

          5.4.1   Prepare  calibration   standards   at   a   minimum  of  five
     concentration levels  for  each  parameter  of  interest  as described 1n
      Paragraph  5.3.

          5.4.2   To  each calibration standard, add  a known constant amount of
      one or more  internal standards, and dilute to volume with hexane or other
      suitable solvent.

          5.4.3   Analyze each calibration  standard according  to  Section 7.0.

      5.5  Surrogate  standards;  The analyst   should monitor the  performance of
the   extraction,cleanup(when   used),    and   analytical   system  and  the
effectiveness of  the method  in dealing with  each  sample matrix by  spiking each
sample,  standard, and  reagent water  blank  with  one or two  surrogates  (e.g.,
organophosphorous pesticides  not  expected to  be  present in   the sample)
recommended  to  encompass the range  of  the temperature program  used in this
method.  Deuterated  analogs  of analytes  should   not be used  as  surrogates for
gas  chromatographic  analysis due  to coelutlon problems.


6.0   SAMPLE  COLLECTION,  PRESERVATION, AND  HANDLING

      6.1   See  the Introductory   material   to this   chapter,  Organic Analytes,
Section  4.1.   Extracts must  be   stored  under refrigeration  and analyzed  within
40 days  of extraction.


 7.0   PROCEDURE

      7.1  Extraction;

           7.1.1  Refer to Chapter Two  for guidance on choosing the appropriate
      extraction procedure.    In  general,   water  samples  are  extracted at a
      neutral,  or as is, pH  with  methylene chloride,  using either Method 3510
      or 3520.   Solid samples are extracted using either Method 3540 or 3550.

           7.1.2  Prior to gas chromatographic analysis, the extraction solvent
      may be exchanged to hexane.  This  is recommended if the detector used is
      halogen-specific.  The exchange  1s  performed  during the K-D procedures
      listed in all  of the   extraction  methods.   The exchange  is performed as
      follows.

                                   8140 -  6
                                                          Revision      0
                                                          Date   September 1986

-------
              7.1.2.1  Following K-D of the methylene chloride extract to
         1 ml using the macro-Snyder column,  allow the apparatus to cool and
         drain for at least  10 min.

              7.1.2.2  Momentarily remove the  Snyder  column,  add  50 ml of
         hexane,  a new boiling  chip,  and  reattach the macro-Snyder column.
         Concentrate the  extract using 1  ml  of  hexane to prewet the Snyder
         column.  Place the   K-D  apparatus  on  the  water  bath so that the
         concentrator tube is partially  immersed  in  the hot water.  Adjust
         the vertical position of the apparatus and the water temperature, as
         required, to complete concentration in 5-10 min.  At the proper rate
         of distillation  the balls of  the  column will actively chatter, but
         the chambers will not  flood.    When  the apparent volume of liquid
         reaches  1 ml, remove the  K-D  apparatus  and  allow it to drain and
         cool for at least 10 min.

              7.1.2.3  Remove the Snyder column   and  rinse the flask and its
         lower joint into the concentrator  tube with  1-2 ml of hexane.  A
         5-mL syringe  is  recommended  for  this operation.  Adjust the extract
         volume   to  10.0 mL.    Stopper   the  concentrator  tube  and store
         refrigerated  at  4*C  if  further   processing  will  not be performed
         immediately.   If the extract will  be stored  longer than two days,  it
         should  be  transferred to  a  Teflon-sealed   screw-cap vial.  Proceed
         with  gas   chromatographic   analysis   if  further  cleanup  is  not
         required.

     7.2 Gas  chromatography  conditions  (Recommended);

         7.2.1   Column la:   Set   helium  carrier  gas  flow at 30 mL/min flow
     rate.   Column temperature is  set  at   150*C for 1 min  and then programmed
     at 25*C/min  to  220'C  and held.

         7.2.2   Column Ib:   Set  nitrogen   carrier gas flow at 30 mL/min flow
     rate.   Column temperature is  set  at   170*C for 2 min  and then programmed
     at 20«C/nrin  to  220*C  and held.

         7.2.3   Column 2:  Set   helium  carrier   gas  at 25 mL/min flow  rate.
     Column temperature is set at   170*C   for  7  min  and then programmed  at
     10*C/m1n  to  250*C  and held.

         7.2.4   Column 3:  Set nitrogen  carrier  gas at 30 mL/min flow  rate.
     Column temperature is set at   100*C   and  then immediately programmed  at
     25'C/min  to  200*C  and held.

     7.3 Calibration;     Refer   to   Method   8000   for   proper   calibration
techniques"!!  Use  Table  1  and especially  Table 2 for guidance  on  selecting the
lowest point on  the calibration curve.

          7.3.1   The procedure for  internal  or   external   calibration may  be
     used.     Refer  to  Method  8000   for   a  description   of  each   of  these
     procedures.
                                  8140 - 7
                                                         Revision      0
                                                         Date  September 1986

-------
         7.3.2  If cleanup Is performed  on  the  samples, the analyst should
    process a series  of  standards  through  the  cleanup procedure and then
    analyze the samples by GC.    This  will confirm elution patterns and the
    absence of interferents from the reagents.

    7.4  Gas chromatographic analysis:

         7,4.1  Refer to Method 8000.    If the internal standard calibration
    technique is used, add 10 uL of  internal standard to the sample prior to
    injection.

         7.4.2  Follow Section 7.6   in  Method  8000  for instructions on the
    analysis sequence,  appropriate  dilutions,  establishing daily retention
    time windows,  and identification criteria.   Include a mid-level standard
    after  each group of 10 samples in the  analysis sequence.

         7.4.3  Examples  of  chromatograms   for  various  organophosphorous
    pesticides are shown  in Figures  1 through 4.

         7.4.4  Record the  sample   volume  injected  and   the  resulting peak
    sizes  (in area units  or peak  heights).

         7.4.5  Using either  the   internal   or  external  calibration  procedure
     (Method 8000), determine  the  identity  and quantity  of each  component peak
     in the sample  chromatogram  which   corresponds   to  the  compounds  used  for
     calibration purposes.   See  Section  7.8 of  Method 8000 for calculation
     equations.

          7.4.6   If peak  detection  and   identification   are  prevented due to
     interferences, the  hexane extract may   undergo  cleanup  using Method 3620.
     The resultant  extract(s)  may  be   analyzed  by   GC directly  or may undergo
     further cleanup to  remove sulfur using Method 3660.

     7.5  Cleanup;

          7.5.1   Proceed  with  Method 3620,   followed  by, if necessary, Method
     3660,  using  the 10-mL hexane  extracts  obtained  from Paragraph 7.1.2.3.

          7.5.2   Following cleanup,  the  extracts should   be  analyzed by GC,  as
     described in  the previous paragraphs and in Method  8000.


8.0  QUALITY CONTROL.

     8.1  Refer  to  Chapter  One   for  specific  quality  control procedures.
Quality control  to validate sample extraction is covered in  Method 3500 and in
the extraction method utilized.   If  extract cleanup was performed, follow  the
QC in Method 3600 and in  the specific cleanup method.

     8.2  Procedures to check  the  GC  system  operation  are found in Method
8000,  Section 8.6.
                                  8140 - 8
                                                         Revision      0
                                                         Date  September 1986

-------
                         Column: 5% SP-2401 on Suptlcoport
                         Ttmptrature: 170°C 7 Minutes, then
                                    tOOC/Minute to 250°C
                         D«ttctor: Phosphorus-Specific Flame Photometric
                    45678
                      RETENTION TIME (MINUTES)
10
11
12
Figure 1. Gas chromatogram of organophosphorus pesticides (Example 1).
                      8140 - 9
                                                Revision       p
                                                Date   September 1986

-------
    Column: 3% SP-2401
    Program: 170°C 7 Minutes, 10°C/Minute
           to250°C
    Ofttctor: Phosphorus/Nitrogen
                 I

                 7654      32
                        RETENTION TIME (MINUTES)
Figure 2. Gas chromatogram of organophosphorus pesticides (Example 2).
                     8140 - 10
                                              Revision       o
                                              Date  September  1986

-------
Column: 15% SE-54 on Gas Chrom Q
Temperature: 100°C Initial, then
            25<>C/Minute to 200°C
Detector: Hall Electrolytic Conductivity-Oxidative Mode
                      7654321
                         RETENTION TIME (MINUTES)
   Figure 3. Gas chromatogram of organophosphorus pesticides (Example 3).
                           8140  - 11
                                                     Revision       0
                                                     Date   September  1986

-------
                Column: 5% SP-2401 on Supelcoport
                Temperature: 170°C 2 Minutes, then 20°C/Minute to 22QOC
                Detector: Phosphorus-Specific Flame Photometric
                 3456
                  RETENTION TIME (MINUTES)
Figure 4. Gas chromatogram of organophosphorus pesticides (Example 4).
                            8140 - 12
                                                     Revision       0	
                                                     Date  September 1986

-------
          8.2.1  Select a representative spike   concentration  for  each  analyte
     to be measured.    The  quality  control  check  sample  concentrate  (Method
     8000,  Section  8.6)  should  contain    each  analyte   in  acetone  at   a
     concentration 1,000  times  more  concentrated   than   the  selected  spike
     concentration.

          8.2.2  Table 3 indicates Single  Operator  Accuracy and Precision for
     this method.  Compare  the  results  obtained  with the  results given  in
     Table 3 to determine if the data quality is acceptable.

     8.3  Calculate surrogate standard  recovery  on  all  samples, blanks, and
spikes.  Determine if  the  recovery  is  within limits (limits established by
performing QC procedures outlined in Method 8000,  Section 8.10).

          8.3.1  If recovery is  not  within  limits, the following procedures
     are required.

               •  Check to  be  sure  there  are  no  errors  in calculations,
                  surrogate solutions  and  internal  standards.   Also,  check
                  instrument performance.

               •  Recalculate the data and/or reanalyze  the extract if any of
                  the above checks reveal a problem.

               •  Reextract and reanalyze the sample  if none of the above are
                  a problem or flag the data as "estimated concentration."

     8.4  GC/MS confirmation;

          8.4.1  GC/MS techniques should  be  judiciously  employed to support
     qualitative identifications made with  this  method.  The GC/MS operating
     conditions and procedures  for  analysis  are  those  specified in Method
     8270.

          8.4.2  When  available,  chemical  ionization  mass  spectra  may be
     employed to aid in  the qualitative identification process.

          8.4.3  Should  these  MS  procedures  fail  to  provide satisfactory
     results, additional steps may  be  taken  before reanalysis.  These steps
     may  include the  use  of  alternate  packed  or  capillary GC columns and
     additional cleanup.


9.0  METHOD PERFORMANCE

     9.1  Single-operator accuracy and  precision  studies  have been conducted
using  spiked wastewater  samples.   The  results of these studies are presented
in Table  3.
                                  8140 - 13
                                                         Revision      0
                                                         Date  September 1986

-------
10.0  REFERENCES

1.  Pressley, T.A. and J.E. Longbottoro, "The Determination of Organophosphorus
Pesticides in Industrial and Municipal Wastewater: Method 614," U.S. EPA/EMSL,
Cincinnati, OH, EPA-600/4-82-004, 1982.

2.  Burke, J.A.,  "Gas  Chromatography  for  Pesticide  Residue Analysis; Some
Practical  Aspects,"    Journal  of  the  Association  of  Official Analytical
Chemists 48, 1037, 1965.

3.  U.S. EPA,  "Analysis of  Volatile  Hazardous Substances by GC/MS: Pesticide
Methods Evaluation," Letter Reports 6,  12A,  and 14, EPA Contract 68-03-2697,
1982.

4.  U.S.   EPA,  "Method   622,   Organophosphorous  Pesticides,"  Environmental
Monitoring and Support  Laboratory, Cincinnati, OH 45268.
                                   8140 - 14
                                                          Revision      0
                                                          Date  September 1986

-------
TABLE 3.  SINGLE-OPERATOR ACCURACY AND PRECISION3
Parameter
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Disulfoton
Ethoprop
Fensulfothion
Fenthion
Merphos
Mevinphos
Naled
Parathion methyl
Phorate
Ronnel
Stirophos
Tokuthion
Trichloronate
Average
recovery
72.7
64.6
98.3
109.0
67.4
67.0
72.1
81.9
100.5
94.1
68.7
120.7
56.5
78.0
96.0
62.7
99.2
66.1
64.6
105.0
Standard
deviation
f V \
\ /
18.8
6.3
5.5
12.7
10.5
6.0
7.7
9.0
4.1
17.1
19.9
7.9
7.8
8.1
5.3
8.9
5.6
5.9
6.8
18.6
Spi ke
range
(ug/L)
21-250
4.9-46
1.0-50.5
25-225
11.9-314
5.6
15.6-517
5.2-92
1.0-51.5
23.9-110
5.3-64
1.0-50
15.5-520
25.8-294
0.5-500
4.9-47
1.0-50
30.3-505
5.3-64
20
Number
of
analyses
17
17
18
17
17
7
16
17
18
17
17
18
16
16
21
17
18
16
17
3
 Information taken from Reference 4.
                                  8140 - 15
                                                         Revision      0
                                                         Date  September 1986

-------
                                          METHOD B14O

                                  ORGANOPHOSPHORUS  PESTICIDES
C
    Q
 7.1.1
      I  Choose
     appropriate
     extraction
      procedure
 (see Chapter 2)
                                                     7.4
   Perform GC
 analysis (see
  Method 8000)
 7.1.2
        Exchange
      i  extract-
  Ion solvent to
        hexane
     during K-D
     procedures
  7.2
     Set gas
  chrometography
   conditions
 Is Identlf lca-
tlon C detection
  prevented  by
   interfei—
     ences
7.5.1]

      Cleanup
   using Method
  3620 and 3360
   If necessary
7

.3
MC
fc
CI
tc
Refer to
thod 8000
ir proper
llbratlon
chnlques
                           7.3.2
                                 Process
                                l a  series
                              of standards
                           through cleanup
                               procedure:
                             analyze by GC
                                     8140 -  16
                                                               Revision       0
                                                               Date  September  1986

-------
                                 METHOD 8141A

               ORGANOPHOSPHORUS  COMPOUNDS  BY  GAS  CHROMATOGRAPHY:
                          CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8141 is a capillary gas chromatographic (GC) method used to
determine the  concentration of organophosphorus  (OP)  compounds.   The fused-
silica, open-tubular columns specified in this method offer improved resolution,
better  selectivity,  increased  sensitivity,  and  faster  analysis  than packed
columns.  The compounds listed in  the table below  can be determined by GC using
capillary  columns with  a  flame  photometric  detector  (FPD)  or  a  nitrogen-
phosphorus detector (NPD).   Triazine herbicides can also be determined with this
method when the NPD is used.  Although performance data are presented for each
of the listed chemicals,  it is unlikely that  all of them could be determined in
a  single  analysis.    This  limitation  results  because  the  chemical  and
chromatographic behavior of  many  of these chemicals can  result in co-elution.
The analyst must  select columns,  detectors and calibration  procedures for the
specific analytes of interest  in  a study.  Any listed  chemical  is a potential
method interference when it is not a  target  analyte.
      Compound Name
                                  8141A - 1
CAS Registry No.
OP Pesticides
Aspon,b
Azinphos-methyl
Azinphos-ethyla
Bolstar (Sulprofos)
Carbophenothion8
Chlorfenvinphos8
Chlorpyrifos
Chlorpyrifos methyl8
Coumaphos
Crotoxyphos8
Demeton-0°
Demeton-Sc
Diazinon
Dichlorofenthiona
Dichlorvos (DDVP)
Dicrotophos8
Dimethoate
Dioxathion8'0
Disulfoton
EPN
Ethion"
Ethoprop
Famphur8
Fenitrothion8
Fensulfothion

3244-90-4
86-50-0
2642-71-9
35400-43-2
786-19-6
470-90-6
2921-88-2
5598-13-0
56-72-4
7700-17-6
8065-48-3
8065-48-3
333-41-5
97-17-6
62-73-7
141-66-2
60-51-5
78-34-2
298-04-4
2104-64-5
563-12-2
13194-48-4
52-85-7
122-14-5
115-90-2
              Revision 1
          September 1994

-------
      Compound Name
CAS Registry No.
      Fonophos8
      Fenthion
      Leptophosa/d
      Malathion
      Merphos0
      Mevinphos6
      Monocrotophos
      Naled
      Parathion, ethyl
      Parathion, methyl
      Phorate
      Phosmet8
      Phosphamidon8
      Ronnel
      Stirophos  (Tetrachlorovinphos)
      Sulfotepp
      TEPPd
      Terbufos8
      Thionazina'b (Zinophos)
      Tokuthionb (Protothiofos)
      Trichlorfon8
      Trichloronateb

Industrial Chemicals
      Hexamethylphosphoramide8 (HMPA)
      Tri-o-cresylphosphatea'd (TOCP)

Triazine Herbicides (NPD only)
      Atrazine8
      Simazine3
   944-22-9
    55-38-9
 21609-90-5
   121-75-5
   150-50-5
  7786-34-7
  6923-22-4
   300-76-5
    56-38-2
   298-00-0
   298-02-2
   732-11-6
 13171-21-6
   299-84-3
 22248-79-9
  3689-24-5
 21646-99-1
 13071-79-9
   297-97-2
 34643-46-4
    52-68-6
   327-98-0
   680-31-9
    78-30-8
  1912-24-9
   122-34-9
      a     This analyte has been evaluated using a 30-m column only.
      b     Production discontinued in the U.S., standard not readily available.
      c     Standards may have multiple components because of oxidation.
      d     Compound is extremely toxic or neurotoxic.
      e     Adjacent major/minor peaks can be observed due to cis/trans isomers.
      1.2   A dual-column/dual-detector approach may be used for the  analysis of
relatively clean  extracts.   Two 15- or 30-m x  0.53-mm  ID fused-silica, open-
tubular columns of different  polarities  are connected to an injection tee and
each is connected to a detector.  Analysts are cautioned  regarding the use of a
dual column configuration when their instrument is subject to mechanical stress,
                                   8141A -  2
              Revision 1
          September 1994

-------
when  many  samples   are  analyzed  over  a  short  time,  or  when  extracts  of
contaminated samples are analyzed.

      1.3   Two detectors can be used for the  listed OP chemicals.  The FPD works
by measuring the emission of phosphorus- or sulfur-containing species.  Detector
performance is  optimized by selecting the proper optical filter and  adjusting the
hydrogen and air flows to the flame.  The NPD is a flame ionization  detector with
a  rubidium  ceramic  flame tip  which  enhances the response  of phosphorus-  and
nitrogen-containing analytes.  The  FPD is more sensitive and more selective, but
is a less common detector in environmental laboratories.

      1.4   Table 1  lists method detection limits  (MDLs) for the target analytes,
using 15-m  columns  and  FPD,  for water  and soil  matrices.   Table 2 lists the
estimated quantitation limits (EQLs) for other matrices.  MDLs and EQLs using 30-
m columns will  be very similar to those obtained from 15-m columns.

      1.5   The use of a  15-m column system has not been fully validated for the
determination  of the  following  compounds.     The   analyst must  demonstrate
chromatographic  resolution  of  all   analytes,  recoveries  of greater than  70
percent, with precision of no more than 15 percent RSD,  before data generated on
the 15-m column system can be reported for these, or any additional, analytes:

      Azinphos-ethyl    Ethion      Phosmet
      Carbophenothion   Famphur     Phosphamidon
      Chlorfenvinphos   HMPA        Terbufos
      Dioxathion        Leptophos   TOCP

      1.6   When Method  8141  is  used to analyze  unfamiliar samples,  compound
identifications should be supported by confirmatory analysis.  Sec. 8.0 provides
gas  chromatograph/mass   spectrometer  (GC/MS)  criteria   appropriate  for  the
qualitative confirmation of compound identifications.

      1.7   This method is restricted to use  by,  or  under the supervision of,
analysts experienced  in  the  use  of capillary gas  chromatography  and  in  the
interpretation  of chromatograms.


2.0   SUMMARY OF METHOD

      2.1   Method 8141 provides gas chromatographic conditions for  the detection
of ppb concentrations of organophosphorus compounds.   Prior to the use of this
method,  appropriate sample preparation techniques must be used.   Water samples
are extracted  at a  neutral  pH with methylene chloride   by  using  a separatory
funnel  (Method  3510)  or a  continuous  liquid-liquid extractor  (Method  3520).
Soxhlet extraction  (Method 3540)  or automated Soxhlet extraction  (Method 3541)
using methylene chloride/acetone (1:1) are used for solid samples.  Both neat and
diluted organic liquids  (Method 3580, Waste Dilution) may be analyzed by direct
injection.   Spiked  samples are used  to  verify the  applicability of the  chosen
extraction technique to  each  new  sample  type.  A gas chromatograph with a flame
photometric  or nitrogen-phosphorus  detector is  used  for  this  multiresidue
procedure.
                                  8141A  - 3                         Revision 1
                                                                September 1994

-------
      2.2   Organophosphorus esters and thioesters can hydrolyze under both acid
and  base   conditions.    Samples  prepared  using acid  and  base  partitioning
procedures are not suitable for analysis by Method 8141.

      2.3   Ultrasonic Extraction  (Method  3550)  is not  an  appropriate sample
preparation  method for  Method  8141  and  should  not  be  used  because  of  the
potential for destruction  of  target analytes  during  the ultrasonic extraction
process.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, and 8000, as well as to Sec. 1.1.

      3.2   The use of Florisil  Cleanup  (Method 3620)  for some of the compounds
in this  method has been demonstrated to yield recoveries less than 85 percent and
is therefore not recommended for all compounds.  Refer to Table 2 of Method 3620
for recoveries of organophosphorus compounds.  Use of an FPD often eliminates the
need for  sample cleanup.   If particular circumstances  demand  the  use  of an
alternative cleanup procedure, the analyst must determine the elution profile and
demonstrate that the recovery of each analyte is not less than 85 percent.

      3.3   The  use of  Gel Permeation Cleanup (GPC)  (Method  3640)  for sample
cleanup has been demonstrated to  yield  recoveries of  less than  85 percent  for
many method  analytes  because they  elute before  bis-(2-ethylhexyl)  phthalate.
Method  3640  is  therefore  not  recommended  for  use with this  method,  unless
analytes of  interest  are  listed  in Method 3640 or  are  demonstrated  to give
greater than 85 percent recovery.

      3.4   Use  of  a  flame photometric detector in  the  phosphorus  mode will
minimize interferences from materials that do not contain  phosphorus or sulfur.
Elemental   sulfur   will   interfere  with   the   determination  of  certain
organophosphorus compounds by flame photometric gas chromatography.   If Method
3660 is  used for sulfur cleanup,  only the tetrabutylammonium (TBA)-sulfite option
should  be employed, since  copper and mercury may destroy OP  pesticides.   The
stability of each  analyte  must  be tested to ensure  that the recovery from  the
TBA-sulfite sulfur cleanup step is not less than 85 percent.

      3.5   A halogen-specific  detector  (i.e.,  electrolytic conductivity or
microcoulometry) is very selective for the halogen-containing compounds and may
be used  for  the determination  of  Chlorpyrifos,  Ronnel,  Coumaphos,  Tokuthion,
Trichloronate,   Dichlorvos,  EPN, Naled, and Stirophos only.   Many  of  the OP
pesticides may  also be detected  by the electron capture detector  (ECD); however,
the ECD is not  as specific as the NPD or FPD.  The ECD should only be used when
previous analyses have demonstrated that interferences  will not adversely effect
quantitation, and that the  detector sensitivity is sufficient to meet regulatory
limits.

      3.6   Certain analytes will coelute,  particularly on 15-m columns (Table
3).  If coelution  is  observed,  analysts should  (1) select  a second  column of
different polarity for confirmation, (2) use 30-m x 0.53-mm columns,  or  (3)  use
0.25- or 0.32-mm  ID  columns.    See Figures 1 through  4 for  combinations of
compounds that do not coelute on  15-m columns.


                                   8141A -  4                        Revision 1
                                                                September  1994

-------
 3.7   The following pairs coeluted on the DB-5/DB-210 30-m column pair:

 DB-5  Terbufos/tri-o-cresyl phosphate
       Naled/Simazine/Atrazine
       Dichlorofenthion/Demeton-0
       Trichloronate/Aspon
       Bolstar/Stirophos/Carbophenothion
       Phosphamidon/Crotoxyphos
       Fensulfothion/EPN

DB-210 Terbufos/tri-o-cresyl phosphate
       Dichlorofenthion/Phosphamidon
       Chlorpyrifos, methyl/Parathion, methyl
       Chlorpyrifos/Parathion, ethyl
       Aspon/Fenthion
       Demeton-0/Dimethoate
       Leptophos/Azinphos-methyl
       EPN/Phosmet
       Famphur/Carbophenothion

 See Table 4 for retention times  of these compounds on 30-m columns.

 3.8   Analytical difficulties encountered for target analytes include:

       3.8.1 Tetraethyl pyrophosphate  (TEPP)  is an  unstable  diphosphate
 which  is  readily  hydrolyzed in  water  and  is  thermally labile  (TEPP
 decomposes at  170°C).   Care  must  be taken  to  minimize loss during  GC
 analysis and during sample preparation.   Identification  of bad  standard
 lots is difficult since the electron  impact (El) mass spectrum of TEPP is
 nearly identical to its major breakdown product, triethyl phosphate.

       3.8.2 The water solubility of Dichlorvos  (DDVP) is 10 g/L  at 20°C,
 and recovery is poor from aqueous solution.

       3.8.3 Naled  is  converted  to  Dichlorvos  (DDVP)   on  column   by
 debromination.   This reaction may  also  occur during  sample workup.  The
 extent of  debromination  will  depend on  the  nature of the  matrix being
 analyzed.   The  analyst  must consider the potential for debromination when
 Naled is to be  determined.

       3.8.4 Trichlorfon rearranges and  is dehydrochlorinated  in  acidic,
 neutral,  or basic media to form  Dichlorvos (DDVP)  and hydrochloric acid.
 If this method  is to be used for the determination of organophosphates in
 the  presence  of  Trichlorfon,  the  analyst  should  be  aware   of  the
 possibility of  rearrangement  to  Dichlorvos to prevent misidentification.

       3.8.5 Demeton (Systox)  is  a  mixture of two  compounds;  0,0-diethyl
 0-[2-(ethylthio)ethyl]phosphorothioate  (Demeton-0) and 0,0-diethyl S-[2-
 (ethylthio)ethyl]phosphorothioate (Demeton-S).   Two peaks are observed in
 all  the   chromatograms corresponding  to  these  two  isomers.    It   is
 recommended  that  the  early  eluting compound   (Demeton-S)  be used  for
 quantisation.
                             8141A - 5                         Revision  1
                                                           September 1994

-------
            3.8.6 Dioxathion is a single-component pesticide.  However, several
      extra peaks are observed in the chromatograms of standards.  These peaks
      appear  to be  the  result  of  spontaneous oxygen-sulfur  isomerization.
      Because  of  this,   Dioxathion  is  not  included  in  composite  standard
      mixtures.

            3.8.7 Merphos (tributyl phosphorotrithioite) is a single-component
      pesticide that  is  readily oxidized to  its phosphorotrithioate (Merphos
      oxone).   Chromatographic analysis  of Merphos almost  always  results two
      peaks  (unoxidized  Merphos elutes  first).   As  the relative  amounts  of
      oxidation  of  the  sample  and  the   standard   are  probably  different,
      quantitation based on the sum of both peaks may be most appropriate.

            3.8.8 Retention times of some analytes, particularly Monocrotophos,
      may increase with  increasing concentrations  in the  injector.   Analysts
      should check for retention time shifts in highly contaminated samples.

            3.8.9 Many  analytes  will   degrade  on   reactive   sites  in  the
      chromatographic system.   Analysts must ensure  that  injectors and splitters
      are  free  from  contamination  and  are  silanized.    Columns   should  be
      installed and maintained properly.

            3.8.10      Performance of chromatographic systems will degrade with
      time.  Column resolution, analyte breakdown and  baselines may be improved
      by column washing (Sec.  7).  Oxidation of columns is not  reversible.

      3.9   Method interferences  may be caused  by contaminants  in  solvents,
reagents, glassware,  and other sample processing hardware that lead to discrete
artifacts or elevated baselines in  gas chromatograms.  All these materials must
be routinely demonstrated to be free  from interferences under the conditions of
the analysis by analyzing reagent blanks  (Sec. 8.0).

      3.10  NP Detector interferences:    Triazine herbicides, such as Atrazine
and Simazine, and other nitrogen-containing compounds may interfere.


4.0   APPARATUS AND MATERIALS

      4.1   Gas  chromatograph:  An  analytical   system   complete  with  a  gas
chromatograph  suitable  for on-column  or  split/splitless  injection, and all
required accessories, including  syringes,  analytical  columns,  gases, suitable
detector(s), and a  recording device.  The  analyst should  select the detector for
the specific measurement application, either the flame photometric detector or
the nitrogen-phosphorus detector.  A data  system for  measuring  peak areas and
dual display of chromatograms  is highly recommended.

            4.1.1 Capillary Columns (0.53-mm,  0.32-mm, or 0.25-mm ID x 15-m or
      30-m length,  depending on the resolution required).  Columns of  0.53-mm ID
      are recommended for most environmental  and waste analysis applications.
      Dual-column, single-injector analysis  requires columns of equal  length and
      bore.  See Sec.  3.0 and  Figures  1 through  4 for  guidance on selecting the
      proper length and diameter for the  column(s).
                                   8141A -  6                         Revision 1
                                                                September 1994

-------
            4.1.1.1     Column  1   -  15-  or  30-m  x  0.53-mm  wide-bore
      capillary column,  1.0-fj.m film thickness,  chemically bonded with 50%
      trifluoropropyl polysiloxane, 50% methyl  polysiloxane  (DB-210), or
      equivalent.

            4.1.1.2     Column  2   -  15-  or  30-m  x  0.53-mm  wide-bore
      capillary column,  0.83-jum film thickness,   chemically bonded with
      35%  phenyl  methyl    polysiloxane  (DB-608,  SPB-608,   RTx-35),  or
      equivalent.

            4.1.1.3     Column  3   -  15-  or  30-m  x  0.53-mm  wide-bore
      capillary column,  1.0 jum film thickness,   chemically bonded with 5%
      phenyl  polysiloxane,  95% methyl   polysiloxane (DB-5, SPB-5, RTx-5),
      or equivalent.

            4.1.1.4     Column 4 - 15-  or 30-m x 0.53-mm ID fused-silica
      open-tubular column,  chemically bonded  with methyl   polysiloxane
      (DB-1, SPB-1, or equivalent), 1.0-/im or  1.5-jim  film thickness.

            4.1.1.5     (optional) Column rinsing kit:  Bonded-phase column
      rinse kit (J&W Scientific, Catalog no. 430-3000 or equivalent).

      4.1.2 Splitter: If a dual-column, single-injector configuration is
used, the open tubular columns should be connected to one  of the following
splitters, or equivalent:

            4.1.2.1     Splitter  1  -  J&W Scientific  press-fit  Y-shaped
      glass 3-way union  splitter (J&W Scientific, Catalog no. 705-0733).

            4.1.2.2     Splitter 2 -  Supelco 8-in  glass  injection tee,
      deactivated (Supelco, Catalog no. 2-3665M).

            4.1.2.3     Splitter   3   -  Restek   Y-shaped   fused-silica
      connector (Restek, Catalog no.  20405).

      4.1.3 Injectors:

            4.1.3.1     Packed column,  1/4-in injector port with hourglass
      liner are recommended for 0.53-mm column.  These injector ports can
      be fitted with splitters (Sec.  4.0)  for dual-column analysis.

            4.1.3.2     Split/splitless capillary  injectors  operated in
      the split mode are required  for 0.25-mm and 0.32-mm columns.

      4.1.4 Detectors:

            4.1.4.1     Flame Photometric Detector (FPD)  operated in the
      phosphorus-specific  mode is  recommended.

            4.1.4.2     Nitrogen-Phosphorus  Detector (NPD) operated  in the
      phosphorus-specific  mode is  less  selective but can  detect  triazine
      herbicides.
                            8141A  - 7                         Revision 1
                                                          September 1994

-------
                  4.1.4.3     Halogen-Specific     Detectors     (electrolytic
            conductivity or  microcoulometry)  may be  used  only  for  a limited
            number of halogenated or sulfur-containing analytes (Sec. 3.0).

                  4.1.4.4     Electron-capture  detectors  may  be  used  for  a
            limited number of analytes (Sec. 3.0).

            4.1.5 Data system:

                  4.1.5.1     Data system capable of presenting chromatograms,
            retention time, and peak integration data is strongly recommended.

                  4.1.5.2     Use of a data system  that allows storage  of raw
            chromatographic data is strongly recommended.


5.0   REAGENTS

      5.1   Solvents

            5.1.1 Isooctane,  (CH3)3CCH2CH(CH3)2 -  Pesticide qual ity or equivalent.

            5.1.2 Hexane, C6H14 - Pesticide quality or equivalent.

            5.1.3 Acetone,  CH3COCH3  -  Pesticide  quality or equivalent.

            5.1.4 Tetrahydrofuran (THF),  C4H80 -  Pesticide quality or equivalent
      (for triazine standards only).

            5.1.5 Methyl tert-butyl-ether (MTBE), CH3Ot-C4H9 - Pesticide quality
      or equivalent (for triazine standards only).

      5.2   Stock standard  solutions  (1000 mg/L):  Can  be prepared  from  pure
standard materials or can be purchased as certified solutions.

            5.2.1 Prepare stock standard  solutions by accurately weighing about
      0.0100 g of pure compounds.  Dissolve the compounds  in  suitable mixtures
      of acetone and hexane  and  dilute to  volume  in a  10-mL  volumetric  flask.
      If  compound  purity  is  96 percent  or greater,  the weight can be  used
      without correction to  calculate the  concentration of the stock standard
      solution.  Commercially prepared stock standard solutions can be used at
      any concentration  if they are  certified  by  the  manufacturer  or by an
      independent source.

            5.2.2 Both Simazine  and Atrazine  have  low  solubilities in hexane.
      If  Simazine  and  Atrazine  standards  are  required,  Atrazine  should  be
      dissolved in MTBE, and  Simazine should  be dissolved  in  acetone/MTBE/THF
      (1:3:1).

            5.2.3 Composite stock standard: This standard can  be prepared from
      individual  stock  solutions.    The  analyst  must demonstrate   that  the
      individual analytes  and common oxidation  products  are  resolved  by the
      chromatographic  system.   For composite stock  standards  containing  less


                                   8141A  -  8                         Revision 1
                                                                September 1994

-------
      than 25 components,  take exactly 1 ml of each individual  stock solution at
      1000 mg/L, add solvent,  and mix the solutions in a  25-mL  volumetric flask.
      For  example,  for a composite containing  20  individual  standards,  the
      resulting concentration of each component in the mixture, after the volume
      is adjusted to  25 ml,  will  be 40 mg/L.  This  composite solution can be
      further diluted  to  obtain the desired  concentrations.   Composite stock
      standards containing more than 25 components are not recommended.

            5.2.4 Store the standard solutions (stock, composite, calibration,
      internal, and surrogate) at 4°C in Teflon-sealed containers in the dark.
      All standard solutions  should be replaced after two months, or sooner if
      routine  QC  (Sec.   8.0)  indicates  a  problem.    Standards  for  easily
      hydrolyzed chemicals including TEPP, Methyl  Parathion, and Merphos should
      be checked every 30 days.

            5.2.5 It  is recommended that  lots of  standards be subdivided and
      stored  in small  vials.   Individual  vials  should be  used  as  working
      standards to minimize the potential for contamination or  hydrolysis of the
      entire lot.

      5.3   Calibration standards   should be  prepared  at a  minimum  of  five
concentrations  by dilution of the  composite  stock standard with  isooctane or
hexane.     The  concentrations  should  correspond  to the expected  range  of
concentrations found in real  samples and should bracket  the linear range of the
detector.  Organophosphorus calibration standards should be  replaced after one
or two months,  or sooner  if   comparison with  check samples or historical  data
indicates that  there  is a problem.   Laboratories  may wish to  prepare separate
calibration solutions for the easily hydrolyzed standards identified above.

      5.4   Internal standard: Internal  standards should only be used on well-
characterized samples by analysts experienced  in the technique. Use of internal
standards  is  complicated  by  co-elution  of  some OP  pesticides  and  by  the
differences in detector response to dissimilar chemicals.

            5.4.1 FPD response for organophosphorus compounds  is enhanced by the
      presence of sulfur atoms bonded to the phosphorus  atom.   It has not been
      established that a thiophosphate can be used as an internal  standard for
      an OP with a different  numbers  of sulfur  atoms (e.g.,  phosphorothioates
      [P=S] as an  internal standard for phosphates  [P04]) or phosphorodithioates
      [P=S2]).

            5.4.2 If internal  standards  are  to be  used,  the analyst must select
      one or more internal standards that are  similar in analytical behavior to
      the compounds  of interest.  The analyst  must further demonstrate that the
      measurement of the internal  standard is not affected by  method or matrix
      interferences.

            5.4.3 When 15-m  columns are used,  it may  be difficult to  fully
      resolve internal standards from target analytes, method  interferences and
      matrix interferences.  The analyst must demonstrate that the measurement
      of  the   internal   standard   is   not    affected   by  method   or   matrix
      interferences.
                                  8141A - 9                         Revision 1
                                                                September 1994

-------
            5.4.4 The following NPD internal standard has been used for a 30-m
      column pair.  Make  a solution of 1000 mg/L of  l-bromo-2-nitrobenzene.  For
      spiking, dilute this  solution to 5 mg/L.  Use  a  spiking volume of 10 /nL/mL
      of extract. The spiking concentration of  the  internal standards should be
      kept constant  for  all  samples and  calibration  standards.   Since its FPD
      response is small,  l-bromo-2-nitrobenzene is not an appropriate internal
      standard for that detector.  No FPD internal  standard is suggested.

      5.5   Surrogate standard spiking solutions - The analyst should monitor the
performance of the extraction, cleanup (when used), and analytical system, and
the effectiveness of the method in dealing with each sample matrix,  by spiking
each  sample,  standard,   and  blank   with  one   or   two   surrogates  (e.g.,
organophosphorus  compounds  not expected  to be  present  in  the  sample).   If
multiple analytes are to be measured, two  surrogates (an early and  a late eluter)
are recommended.   Deuterated analogs of analytes are not appropriate surrogates
for gas chromatographic/FPD/NPD analysis.

            5.5.1 If surrogates are to be used, the analyst must select one or
      more compounds that are similar in analytical  behavior to the compounds of
      interest.  The analyst must  further demonstrate that  the measurement of a
      surrogate  is  not  affected by method  or  matrix  interferences.   General
      guidance on the selection and use of surrogates is provided  in Sec. 5.0 of
      Method 3500.

            5.5.2 Tributyl  phosphate and  triphenyl phosphate are used  as FPD and
      NPD surrogates.  A  volume of 1.0 ml of  a  1-M9/L spiking solution (1 ng of
      surrogate)  is  added to each water  sample and each  soil/sediment sample.
      If there is a  co-elution problem,  4-chloro-3-nitrobenzo-trifluoride has
      also been used as an NPD-only surrogate.


6.0   SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

      6.1   See the  introductory material to Chapter Four,  "Organic Analytes,"
Sec. 4.0.

      6.2   Extracts are to  be refrigerated at 4°C and analyzed within 40 days
of extraction.  See  Sec.  5.2.4 for storage of standards.

      6.3   Organophosphorus  esters   will  hydrolyze   under  acidic  or  basic
conditions.  Adjust  samples  to a pH of 5 to 8 using  sodium hydroxide or sulfuric
acid solution as  soon as possible  after  sample collection.  Record the volume
used.

      6.4   Even  with storage at  4°C   and  use  of  mercuric chloride   as  a
preservative,  most   OPs  in  groundwater   samples  collected  for   the  national
pesticide survey degraded within a 14-day  period.  Begin sample extraction within
7 days of collection.
                                  8141A - 10                        Revision 1
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Extraction and cleanup:

            7.1.1 Refer to Chapter Two  and Method 8140 for guidance on choosing
      the appropriate  extraction  procedure.   In general,  water  samples  are
      extracted at a  neutral  pH with methylene chloride, using  either Method
      3510 or 3520.   Solid samples are extracted using either Method  3540 or
      3541 with methylene chloride/acetone (1:1 v/v)  or hexane/acetone (1:1 v/v)
      as the extraction  solvent.   Method 3550 is an  inappropriate extraction
      technique for the target analytes of this method (See  Sec.  2.3).

            7.1.2 Extraction and cleanup  procedures  that use solutions below pH
      4 or above pH 8 are not appropriate for this method.

            7.1.3 If required, the samples may be cleaned up using the  Methods
      presented in Chapter Four,  Sec. 2.  Florisil Column Cleanup (Method 3620)
      and Sulfur Cleanup (Method 3660,  TBA-sulfite option) may have particular
      application for  OPs.   Gel  Permeation  Cleanup (Method 3640)  should  not
      generally be used for OP pesticides.

                  7.1.3.1     If sulfur cleanup by Method 3660 is required, do
            not use mercury or copper.

                  7.1.3.2     GPC  may  only   be  employed  if  all  target  OP
            pesticides  are listed  as analytes  of Method   3640,   or  if  the
            laboratory has demonstrated a recovery  of greater  than 85  percent
            for target  OPs at a  concentration not greater  than 5 times  the
            regulatory   action   level.     Laboratories  must   retain   data
            demonstrating acceptable recovery.

            7.1.4 Prior to gas chromatographic  analysis, the  extraction  solvent
      may be  exchanged  to hexane.  The analyst  must ensure quantitative transfer
      of the  extract  concentrate.   Single-laboratory data indicate that  samples
      should  not  be transferred with 100-percent hexane during sample workup, as
      the more  polar  organophosphorus compounds may  be lost.    Transfer  of
      organophosphorus esters is best accomplished using methylene chloride or
      a hexane/acetone solvent mixture.

            7.1.5 Methylene chloride may  be used as an  injection solvent with
      both the  FPD and the NPD.

            NOTE: Follow manufacturer's instructions as to suitability of using
                  methylene chloride with any specific  detector.

      7.2   Gas  chromatographic  conditions:

            7.2.1 Four  0.53-mm  ID  capillary   columns  are   suggested  for  the
      determination of organophosphates  by  this  method.  Column 1  (DB-210 or
      equivalent) and  Column 2  (SPB-608 or  equivalent)  of  30-m  length  are
      recommended if  a large number  of  organophosphorus analytes  are  to  be
      determined.   If superior chromatographic  resolution is  not required, 15-m
      lengths columns may be appropriate.  Operating  conditions for 15-m  columns


                                  8141A -  11                         Revision 1
                                                               September 1994

-------
      are listed in Table 5.  Operating conditions for 30-m columns are listed
      in Table 6.

            7.2.2 Retention  times  for  analytes on  each set  of  columns  are
      presented in Tables 3 and 4.

      7.3   Calibration: Refer to Method 8000 for proper calibration techniques.
Use Table 5 and Table 6 for establishing the proper  operating parameters for the
set of columns being employed in the analyses.

      7.4   Gas chromatographic analysis:   Method 8000 provides instructions on
the analysis sequence,  appropriate  dilutions,  establishing daily retention time
windows and identification criteria.

            7.4.1 Automatic injections  of 1 juL are recommended.  Hand injections
      of no more than  2 /LtL may be used  if the  analyst demonstrates quantitation
      precision of < 10 percent relative standard deviation.  The solvent flush
      technique may be used if the amount of solvent is kept at a minimum.   If
      the internal  standard calibration technique is used, add 10 juL of internal
      standard to each ml  of  sample  prior  to  injection.   Chromatograms of the
      target organophosphorus compounds are shown in Figures 1  through 4.

            7.4.2 Figures 5 and 6 show  chromatograms with and without Simazine,
      Atrazine, and Carbophenothion on 30-m columns.

      7.5   Record the  sample  volume injected to the nearest 0.05 pi and  the
resulting peak sizes  (in  area units or peak heights).  Using either the internal
or external  calibration procedure  (Method  8000),  determine the  identity  and
quantity of each component peak in  the  sample  chromatogram which corresponds to
the compounds  used for  calibration  purposes.   See  Method 8000  for calculation
equations.

            7.5.1 If  peak detection and  identification  is  prevented  by  the
      presence of interferences, the use of an FPD or further sample cleanup is
      required.  Before using any cleanup procedure, the  analyst must process a
      series of calibration standards through the procedure to establish elution
      patterns and to  determine recovery of target  compounds.   The absence of
      interference from reagents must be demonstrated by routine processing of
      reagent  blanks through the chosen cleanup procedure.   Refer to Sec.  3.0
      for interferences.

            7.5.2 If the responses  exceed the linear range of the system, dilute
      the extract and reanalyze.  It  is recommended that  extracts be diluted so
      that  all peaks  are on scale.  Overlapping peaks  are  not  always evident
      when  peaks  are  off-scale.    Computer   reproduction  of  chromatograms,
      manipulated to ensure all peaks  are  on  scale  over a 100-fold range,  are
      acceptable if linearity  is  demonstrated.  Peak  height measurements  are
      recommended over peak area integration when overlapping peaks cause errors
      in area  integration.

            7.5.3 If the peak response  is less than 2.5  times the baseline noise
      level, the validity of the quantitative result may be questionable.  The
                                  8141A - 12                        Revision 1
                                                                September 1994

-------
      analyst should consult with  the  source of the sample to determine whether
      further concentration of the sample extract is  warranted.

            7.5.4 If partially overlapping  or coeluting peaks are found, change
      columns or try a GC/MS technique.   Refer to Sec.  8.0 and Method 8270.

      7.6   Suggested chromatograph maintenance: Corrective measures may require
any one or more of the following remedial actions.

            7.6.1 Refer  to  Method   8000  for   general   information  on  the
      maintenance of capillary columns and injectors.

            7.6.2 Splitter  connections:  For  dual  columns which  are  connected
      using  a  press-fit Y-shaped glass  splitter or a  Y-shaped  fused-silica
      connector  (J&W Scientific,  Restek, or  equivalent),  clean  and  deactivate
      the splitter.   Reattach the columns after cleanly  cutting off at least one
      foot from the  injection port side of  the column using a capillary cutting
      tool or  scribe.   The  accumulation of  high boiling residues can change
      split ratios between  dual  columns and thereby change calibration factors.

            7.6.3 Columns  will  be  damaged  permanently   and  irreversibly  by
      contact with oxygen at elevated  temperature.  Oxygen can enter the column
      during a septum change,  when oxygen traps are exhausted, through neoprene
      diaphragms of regulators, and through leaks in the  gas  manifold.   Polar
      columns  including  the DB-210 and  DB-608 are  more prone  to  oxidation.
      Oxidized  columns  will  exhibit  baselines  that  rise  rapidly  during
      temperature programming.

            7.6.4 Peak tailing for all components will  be exacerbated by dirty
      injectors, pre-columns, and glass  "Y"s.  Additionally,  cleaning  of this
      equipment  (or replacement/clipping,  as appropriate)  will  greatly reduce
      the peak  tailing.    Components  such  as Fensulfothion,  Naled,  Azinphos-
      methyl, and Dimethoate are very  good indicators of  system performance.

      7.7   Detector maintenance:

            7.7.1 Older  FPDs  may  be  susceptible   to  stray  light  in  the
      photomultiplier tube  compartment.   This stray light will decrease  the
      sensitivity and the  linearity of  the detector.  Analysts  can  check  for
      leaks by initiating an analysis  in  a  dark room and turning on the lights.
      A shift  in the baseline  indicates  that  light may  be leaking  into  the
      photomultiplier tube  compartment.  Additional shielding should be applied
      to eliminate light leaks and minimize stray light interference.

            7.7.2 The bead  of  the NPD will become exhausted with time,  which
      will decrease the sensitivity and the selectivity  of  the  detector.   The
      collector may become  contaminated  which decreased detector sensitivity.

            7.7.3 Both types of detectors  use a  flame  to  generate  a  response.
      Flow rates  of air and  hydrogen should  be optimized  to  give  the most
      sensitive, linear detector response for target  analytes.
                                  8141A -  13                         Revision  1
                                                                September 1994

-------
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality  control  procedures.
Include a mid-level check standard after each group of 10 samples  in the analysis
sequence.  Quality control to  validate  sample  extraction  is  covered in Method
3500 and in the extraction method utilized.   If extract cleanup was performed,
follow the QC in Method 3600 and in the specific cleanup method.

      8.2   Procedures to check the GC system operation are found  in Method 8000.

      8.3   GC/MS confirmation

            8.3.1 GC/MS techniques  should  be  judiciously employed  to  support
      qualitative  identifications  made with  this  method.   Follow the  GC/MS
      operating requirements specified in Method 8270.

            8.3.2 When  available,   chemical  ionization  mass spectra  may  be
      employed to aid in the qualitative identification process.

            8.3.3 To confirm an  identification  of  a  compound,  the  background-
      corrected mass; spectrum of the compound  must be obtained from the sample
      extract  and  must  be  compared  with  a  mass spectrum  from  a stock  or
      calibration standard analyzed under the  same chromatographic  conditions.
      At  least  25 ng  of material  should  be  injected into  the  GC/MS.    The
      following criteria must be met for qualitative  confirmation:

                  8.3.3.1     The   qualitative   identification   of  compounds
            determined  by  this  method  is  based  on  retention  time,  and  on
            comparison of the sample mass spectrum,  after background correction,
            with  characteristic  ions   in  a  reference mass spectrum.    The
            reference mass spectrum must be generated  by  the laboratory  using
            the conditions of  this  method.   The characteristic  ions  from the
            reference mass spectrum  are defined to be  the three ions of greatest
            relative intensity, or any  ions over  30% relative intensity if less
            than three  such  ions occur in  the  reference  spectrum.   Compounds
            should be identified as present when the  criteria below are met.

                        8.3.3.1.1   The intensities of the characteristic ions
                  of a compound maximize in the  same scan or within one scan of
                  each  other.   Selection of  a peak by  a data  system target
                  compound search  routine  where  the  search  is based on  the
                  presence of  a target  chromatographic  peak containing  ions
                  specific  for  the target  compound  at  a  compound-specific
                  retention  time will be accepted as  meeting this criterion.

                        8.3.3.1.2   The RRT of  the sample  component is within
                  ± 0.06 RRT units of the RRT of the  standard component.

                        8.3.3.1.3   The    relative    intensities    of    the
                  characteristic  ions   agree   within   30%   of   the  relative
                  intensities  of   these  ions   in   the   reference  spectrum.
                  (Example:     For  an  ion  with  an abundance  of  50% in  the
                  reference  spectrum, the corresponding abundance  in  a sample
                  spectrum can range between 20% and  80%.)

                                  8141A - 14                        Revision 1
                                                                September 1994

-------
            8.3.3.1.4   Structural isomers that produce very similar
      mass  spectra  should be identified  as  individual  isomers if
      they  have   sufficiently   different   GC   retention  times.
      Sufficient  GC  resolution  is  achieved  if the  height  of the
      valley between two isomer  peaks is less than 25% of the sum of
      the  two peak  heights.    Otherwise,  structural  isomers are
      identified  as  isomeric pairs.

            8.3.3.1.5   Identification  is  hampered  when  sample
      components  are not  resolved chromatographically and produce
      mass  spectra  containing  ions contributed  by  more  than one
      analyte.  When gas chromatographic peaks obviously represent
      more  than one  sample  component  (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of  analyte  spectra  and  background
      spectra  is  important.   Examination  of extracted ion current
      profiles  of appropriate  ions  can  aid  in  the  selection of
      spectra, and in qualitative identification of compounds.  When
      analytes  coelute (i.e.,  only one  chromatographic peak is
      apparent),  the  identification criteria  can  be  met, but each
      analyte  spectrum will  contain extraneous ions contributed by
      the coeluting  compound.

      8.3.3.2     For samples containing components not associated
with the calibration  standards,  a library search may be made for the
purpose of tentative identification.  The  necessity to perform this
type of  identification will be determined  by the purpose  of the
analyses  being  conducted.    Computer  generated  library  search
routines   should   not  use   normalization   routines  that   would
misrepresent the  library  or unknown spectra  when  compared to each
other.   For example,  the RCRA permit or waste delisting requirements
may require the reporting  of nontarget analytes.  Only after visual
comparison of sample spectra with  the nearest  library searches will
the  mass spectral  interpretation specialist  assign  a  tentative
identification. Guidelines for making tentative identification are:

      (1)   Relative  intensities  of major  ions  in  the  reference
spectrum (ions > 10% of the  most abundant  ion) should be present in
the sample spectrum.

      (2)   The relative intensities of  the major ions should agree
within ± 20%.   (Example:  For an ion with an  abundance of  50% in the
standard spectrum, the corresponding sample  ion  abundance must be
between 30 and 70%.)

      (3)   Molecular ions present in the reference spectrum should
be present in the sample spectrum.

      (4)   Ions  present  in the  sample spectrum  but not  in  the
reference  spectrum  should  be  reviewed  for  possible  background
contamination or presence of coeluting  compounds.

      (5)   Ions present in the reference spectrum but not  in the
sample  spectrum should be reviewed for possible subtraction from the

                      8141A  - 15                        Revision 1
                                                    September 1994

-------
            sample spectrum  because  of background  contamination  or coeluting
            peaks.  Data system library reduction programs can sometimes create
            these discrepancies.

            8.3.4 Where  available,  chemical  ionization mass  spectra  may  be
      employed to aid in the qualitative identification process because of the
      extensive fragmentation  of organophosphorus  pesticides  during electron
      impact MS processes.

            8.3.5 Should the MS procedure  fail to provide satisfactory results,
      additional steps may be taken before reanalysis.   These steps may  include
      the use of alternate packed or capillary GC columns or additional sample
      cleanup.


9.0   METHOD PERFORMANCE

      9.1   Estimated MDLs and associated chromatographic conditions for water
and clean soil (uncontaminated with synthetic organics) are listed in Table 1.
As detection limits will vary with the particular matrix to be analyzed, guidance
for determining EQLs  is given in Table 2.  Recoveries for several method analytes
are provided in Tables 5, 6, and 7.


10.0  REFERENCES

1.    Taylor, V.; Mickey, D.M.; Marsden, P.J. "Single Laboratory Validation of
      EPA Method  8140"; U.S.  Environmental  Protection  Agency,  Environmental
      Monitoring Systems  Laboratory,  Office of Research and  Development,  Las
      Vegas, NV, 1987; EPA-600/4-87-009.

2.    Pressley, T.A;  Longbottom,  J.E.   "The  Determination  of Organophosphorus
      Pesticides  in  Industrial  and  Municipal Wastewater:  Method  614";  U.S.
      Environmental   Protection Agency,  Environmental  Monitoring  and   Support
      Laboratory, Cincinnati, OH, 1982; EPA-600/4-82-004.

3.    "Analysis of Volatile  Hazardous  Substances  by  GC/MS:  Pesticide  Methods
      Evaluation"; Letter  Reports 6,  12A,  and  14 to  the  U.S.  Environmental
      Protection Agency on Contract 68-03-2697,  1982.

4.    "Method 622, Organophosphorus Pesticides";  U.S. Environmental  Protection
      Agency, Environmental  Monitoring  and  Support Laboratory,  Cincinnati,  OH
      45268.

5.    Lopez-Avila, V.;  Baldin,  E.;  Benedicto, J;  Milanes, J.;  Beckert, W.  F.
      "Application of Open-Tubular Columns to SW-846 GC Methods"; final report
      to the U.S. Environmental Protection Agency on Contract 68-03-3511; Mid-
      Pacific Environmental  Laboratory, Mountain View,  CA,  1990.

6.    Hatcher,  M.D.;   Hickey, D.M.;   Marsden,  P.J.;  and    Betowski, L.D.;
      "Development of a GC/MS Module  for RCRA Method 8141"; final report to the
      U.S. EPA Environmental  Protection Agency on Contract 68-03-1958; S-Cubed,
      San Diego, CA, 1988.


                                  8141A - 16                         Revision  1
                                                                September 1994

-------
7.    Chau, A.S.Y.; Afghan, B.K.  Analysis  of Pesticides  in Water; "Chlorine and
      Phosphorus-Containing Pesticides"; CRC:  Boca Raton, FL, 1982, Vol. 2, pp
      91-113, 238.

8.    Hild,  J.;  Schulte,   E;  Thier,   H.P.  "Separation  of  Organophosphorus
      Pesticides   and   Their    Metabolites   on   Glass-Capillary   Columns";
      Chromatographia, 1978, 11-17.

9.    Luke,  M.A.;  Froberg,  J.E.;  Doose,  G.M.;  Masumoto,  H.T.  "Improved
      Multiresidue  Gas  Chromatographic  Determination  of  Organophosphorus,
      Organonitrogen,  and  Organohalogen  Pesticides  in  Produce,   Using  Flame
      Photometric and Electrolytic  Conductivity Detectors"; J. Assoc. Off, Anal.
      Chem. 1981, 1187, 64.

10.   Sherma, J.;  Berzoa,   M.  "Analysis  of  Pesticide  Residues  in Human  and
      Environmental  Samples";  U.S. Environmental  Protection Agency,  Research
      Triangle Park, NC; EPA-600/8-80-038.

11.   Desmarchelier,  J.M.;  Wustner,   D.A.;  Fukuto,  T.R.  "Mass  Spectra  of
      Organophosphorus Esters and Their Alteration Products"; Residue Reviews,
      1974, pp 63, 77.

12.   Munch, D.J. and Frebis, C.P., "Analyte Stability Studies Conducted during
      the National Pesticide Survey",  ES & 7",  1992,  vol  26, 921-925.

13.   T.L. Jones, "Organophosphorus Pesticide Standards:  Stability Study", EMSL-
      LV Research Report, EPA 600/X-92/040, April, 1992

14.   Kotronarou, A., et  al., "Decomposition of Parathion in Aqueous Solution by
      Ultrasonic Irradiation,"  £5*7,  1992, Vol.  26,  1460-1462.
                                  8141A - 17                        Revision 1
                                                                September 1994

-------
                              TABLE  1
           METHOD  DETECTION  LIMITS  IN  A  WATER  AND  A  SOIL
    MATRIX USING 15-m COLUMNS AND A FLAME PHOTOMETRIC DETECTOR
Compound
Reagent
Water (3510)'
(M9/L)
                                                      Soil (3540)
Azinphos-methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton, -0, -S
Diazinon
Dichlorvos (DDVP)
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotepp
TEPP°
Tetrachlorovinphos
Tokuthion (Protothiofos)c
Trichloronate0
0.10
0.07
0.07
0.20
0.12
0.20
0.80
0.26
0.07
0.04
0.20
0.08
0.08
0.11
0.20
0.50
0.50
0.06
0.12
0.04
0.07
0.07
0.80
0.80
0.07
0.80
5.0
3.5
5.0
10.0
6.0
10.0
40.0
13.0
3.5
2.0
10.0
4.0
5.0
5.5
10.0
25.0
25.0
3.0
6.0
2.0
3.5
3.5
40.0
40.0
5.5
40.0
Sample  extracted  using  Method  3510,  Separatory  Funnel  Liquid-Liquid
Extraction.

Sample extracted using Method 3540, Soxhlet Extraction.

Purity  of these  standards  not  established  by  the EPA  Pesticides  and
Industrial Chemicals Repository, Research Triangle  Park, NC.
                            8141A - 18
                                Revision 1
                            September 1994

-------
                                    TABLE 2
                DETERMINATION OF ESTIMATED QUANTITATION LIMITS
                          (EQLs)  FOR VARIOUS MATRICES8
   Matrix                                                            Factor
   Ground water (Methods 3510 or 3520)                                 10b
   Low-concentration soil by Soxhlet and no cleanup                    10°
   Non-water miscible waste (Method 3580)                            1000
c
8  EQL = [Method detection limit (see Table 1)] X  [Factor found in this table].
For non-aqueous samples, the factor  is on a wet-weight basis.  Sample EQLs are
highly matrix dependent.  The EQLs to be  determined herein are for guidance and
may not always be achievable.

b  Multiply this factor times the reagent water MDL in Table 1.

c  Multiply this factor times the soil MDL in Table 1.
                                  8141A - 19                        Revision 1
                                                                September 1994

-------
                TABLE 3.
RETENTION TIMES FOR METHOD 8141A ANALYTES
          EMPLOYING  15-m COLUMNS


TEPP
Dichlorvos (DDVP)
Mevinphos
Demeton, -0 and -S
Ethoprop
Naled
Phorate
Monochrotophos
Sulfotepp
Dimethoate
Disulfoton
Diazinon
Merphos
Ronnel
Chlorpyrifos
Malathion
Parathion, methyl
Parathion, ethyl
Trichloronate
Tetrachlorovinphos
Tokuthion (Protothiofos)
Fensulfothion
Bolstar^ (Sulprofos)
Famphur*
EPN
Azinphos-methyl
Fenthion
Coumaphos
"Method 8141A has not been fully
Initial temperature
Initial time
Program 1 rate
Program 1 final temp.
Program 1 hold
Program 2 rate
Program 2 final temp.
Program 2 hold
Caoil
Compound

9.63
14.18
18.31
18.62

19.94
20.04
20.11
20.64
23.71
24.27
26.82
29.23
31.17
31.72
31.84
31.85
32.19
34.65
34.67
35.85
36.34
36.40

38.34
38.83
39.83
lary Column
OB-5
6.44
7.91
12.88
15.90
16.48
19.01
17.52
20.11
18.02
20.18
19.96
20.02
21.73
22.98
26.88
28.78
23.71
27.62
28.41
32.99
24.58
35.20
35.08
36.93
37.80
38.04
29.45
38.87

SPB-608
5.12
12.79
18.44
17.24
18.67
17.40
18.19
31.42
19.58
27.96
20.66
19.68
32.44
23.19
25.18
32.58
32.17
33.39
29.95
33.68
39.91
36.80
37.55
37.86
36.71
37.24
28.86
39.47

DB-210
10.66




19.35


















36.74



validated for Famphur.
130°C
3 minutes
5°C/min
180°C
10 minutes
2°C/min
250°C
15 minutes
50°C
1 minute
5°C/min
140°C
10 minutes
10°C/min
240°C
10 minutes
50°C
1 minute
5°C/min
140°C
10 minutes
10°C/min
240°C
10 minutes








             8141A  -  20
    Revision 1
September 1994

-------
                                   TABLE 4.
                   RETENTION TIMES FOR METHOD 8141A ANALYTES
                            EMPLOYING 30-m COLUMNS8
Compound
DB-5
      RT (min)

DB-210      DB-608
DB-1
Trimethyl phosphate
Dichlorvos (DDVP)
Hexamethyl phosphorami de
Trichlorfon
TEPP
Thionazin
Mevinphos
Ethoprop
Diazinon
Sulfotepp
Terbufos
Tri-o-cresyl phosphate
Naled
Phorate
Fonophos
Disulfoton
Merphos
Oxidized Merphos
Dichlorofenthion
Chlorpyrifos, methyl
Ronnel
Chlorpyrifos
Trichloronate
Aspon
Fenthion
Demeton-S
Demeton-0
Monocrotophos0
Dimethoate
Tokuthion
Malathion
Parathion, methyl
Fenithrothion
Chlorfenvinphos
Parathion, ethyl
Bolstar
Stirophos
Ethion
b
7.45
b
11.22
b
12.32
12.20
12.57
13.23
13.39
13.69
13.69
14.18
12.27
14.44
14.74
14.89
20.25
15.55
15.94
16.30
17.06
17.29
17.29
17.87
11.10
15.57
19.08
18.11
19.29
19.83
20.15
20.63
21.07
21.38
22.09
22.06
22.55
2.36
6.99
7.97
11.63
13.82
24.71
10.82
15.29
18.60
16.32
18.23
18.23
15.85
16.57
18.38
18.84
23.22
24.87
20.09
20.45
21.01
22.22
22.73
21.98
22.11
14.86
17.21
15.98
17.21
24.77
21.75
20.45
21.42
23.66
22.22
27.57
24.63
27.12

6.56

12.69


11.85
18.69
24.03
20.04
22.97

18.92
20.12

23.89

35.16
26.11
26.29
27.33
29.48
30.44

29.14
21.40
17.70
19.62
20.59
33.30
28.87
25.98

32.05
29.29
38.10
33.40
37.61

10.43




14.45
18.52
21.87
19.60


18.78
19.65

21.73
26.23



23.67
24.85


24.63
20.18

19.3
19.87
27.63
24.57
22.97


24.82
29.53
26.90

                                                                   (continued)
                                  8141A - 21
                            Revision 1
                        September 1994

-------
                             TABLE 4. (Continued)
Compound
 DB-5
       RT  (min)

 DB-210      DB-608
           DB-1
Phosphamidon
Crotoxyphos
Leptophos
Fensulfothion
EPN
Phosmet
Azinphos-methyl
Azinphos-ethyl
Famphur
Coumaphos
Atrazine
Simazine
Carbophenothion
Dioxathion
Trithion methyl
Dicrotophos
Internal Standard
1-Bromo-2-nitrobenzene
Surrogates
Tributyl phosphate
Triphenyl phosphate
4-Cl-3-nitrobenzotrifluoride
22.77
22.77
24.62
27.54
27.58
27.89
28.70
29.27
29.41
33.22
13.98
13.85
22.14
 d
 8.11
5.73
20.09
23.85
31.32
26.76
29.99
29.89
31.25
32.36
27.79
33.64
17.63
17.41
27.92
 9.07
 5.40
25.88
32.65
44.32
36.58
41.94
41.24
43.33
45.55
38.24
48.02
              22.24
              36.62
              19.33
                       11.1
                       33.4
28.58
31.60

32.33
34.82
a The GC operating conditions were as follows:

DB-5 and DB-210  - 30-m  x  0.53-mm  ID  column,  DB-5 (1.50-  m film thickness) and
DB-210 (1.0- m film thickness).  Both connected to a  press-fit Y-shaped inlet
splitter.   Temperature  program:  120°C  (3-min hold) to 270°C  (10-min  hold)  at
5°C/min; injector temperature 250°C; detector temperature 300°C; bead temperature
400°C; bias voltage  4.0;  hydrogen gas pressure  20 psi;  helium  carrier  gas 6
mL/min; helium makeup gas 20 mL/min.

DB-608 - 30-m x 0.53-mm ID column,  DB-608  (1.50-  m  film thickness) installed in
an 0.25-in  packed-column  inlet.   Temperature program:  110°C (0.5-min  hold)  to
250°C (4-min hold) at 3°C/min;  injector temperature 250°C; helium carrier gas 5
mL/min; flame photometric detector.

DB-1 30-m x  0.32-mm ID column, DB-1  (0.25- m film thickness) split/splitless with
head pressure of  10  psi,  split  valve closure at 45 sec,  injector temp. 250°C,
50°C (1-min hold) to 280°C (2-min  hold) at 6°C/min, mass spectrometer full scan
35-550 amu.

b Not detected at 20 ng per  injection.
c Retention times may shift to longer times with larger amounts  injected (shifts
of over 30  seconds have been observed,  Hatcher et. al.)
d Shows multiple  peaks; therefore, not included  in the composite.
                                  8141A - 22
                              Revision 1
                          September 1994

-------
                               TABLE  5.
PERCENT RECOVERY OF 27 ORGANOPHOSPHATES BY SEPARATORY FUNNEL EXTRACTION
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate

Low
126
134
7
103
33
136
80
NR
48
113
82
84
NR
127
NR
NR
NR
NR
101
NR
94
67
87
96
79
NR
NR
Percent Recovery
Medium
143 + 8
141 + 8
89 + 6
90 + 6
67 + 11
121 + 9.5
79 + 11
47 + 3
92 + 7
125 + 9
90 + 6
82 + 12
48 + 10
92 + 6
79
NR
18 + 4
NR
94 + 5
46 + 4
77 + 6
97 + 5
85 + 4
55 + 72
90 + 7
45 + 3
35

High
101
101
86
96
74
82
72
101
84
97
80
96
89
86
81
55
NR
NR
86
44
73
87
83
63
80
90
94
NR = Not recovered.
                            8141A - 23
    Revision 1
September 1994

-------
                                   TABLE 6.
PERCENT RECOVERY OF 27 ORGANOPHOSPHATES BY CONTINUOUS LIQUID-LIQUID EXTRACTION
Percent Recovery
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Famphur
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
NR
NR
13
94
38
NR
81
NR
94
NR
39
_.
90
8
105
NR
NR
NR
NR
106
NR
84
82
40
39
56
132
NR
Medium
129
126
82 + 4
79 + 1
23 + 3
128 + 37
32 + 1
10 + 8
69 + 5
104 + 18
76 + 2
63 + 15
67 + 26
32 + 2
87 + 4
80
87
30
NR
81 + 1
50 + 30
63 + 3
83 + 7
77 + 1
18 + 7
70 + 14
32 + 14
NR
High
122
128
88
89
41
118
74
102
81
119
83
--
90
86
86
79
49
1
74
87
43
74
89
85
70
83
90
21
    NR = Not recovered.
                                8141A - 24
    Revision 1
September 1994

-------
                               TABLE 7.
     PERCENT RECOVERY OF 27 ORGANOPHOSPHATES BY  SOXHLET EXTRACTION
Percent Recovery
Compound
Azinphos methyl
Bolstar
Chlorpyrifos
Coumaphos
Demeton
Diazinon
Dichlorvos
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfonthion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Sulfotep
TEPP
Tetrachlorvinphos
Tokuthion
Trichloroate
Low
156
102
NR
93
169
87
84
NR
78
114
65
72
NR
100
62
NR
NR
NR
75
NR
75
NR
67
36
50
NR
56
Medium
110 + 6
103 + 15
66 + 17
89 + 11
64 + 6
96 + 3
39 + 21
48 + 7
78 + 6
93 + 8
70 + 7
81 + 18
43 + 7
81+8
53
71
NR
48
80 + 8
41 + 3
77 + 6
83 + 12
72 + 8
34 + 33
81 + 7
40 + 6
53
High
87
79
79
90
75
75
71
98
76
82
75
111
89
81
60
63
NR
NR
80
28
78
79
78
63
83
89
53
NR = Not recovered,
                            8141A - 25
    Revision 1
September 1994

-------
                               TABLE 8.

            SUGGESTED OPERATING CONDITIONS FOR 15-m COLUMNS
Columns 1 and 2 (DB-210 and SPB-608 or their equivalent)
Carrier gas (He) flow rate =
Initial temperature =
Temperature program =
Column 3  (DB-5 or equivalent)

Carrier gas (He) flow rate =
Initial temperature =
Temperature program =
5 mL/min
50°C, hold for 1 minute
50'C to 140°C at 5°C/min, hold for
10 minutes,  followed by  140°C  to
240°C  at  10°C/min,   hold  for  10
minutes (or a sufficient amount of
time for last compound to elute).
5 mL/min
130°C, hold for 3 minutes
130°C to 180°C at  5°C/min, hold for
10 minutes,  followed by  180'C  to
250°C  at  2°C/min,  hold  for  15
minutes (or a sufficient amount of
time for last compound to elute).
                             8141A - 26
                    Revision 1
                September  1994

-------
                                      TABLE  9
                  SUGGESTED OPERATING  CONDITIONS  FOR  30-m  COLUMNS
Column 1:
      Type:   DB-210
      Dimensions:   30-m x 0.53-mm ID
      Film Thickness (urn) :  1.0
Column 2:
      Type:   DB-5
      Dimensions:  30-m x 0.53-mm ID
      Film Thickness (nm):  1.5
Carrier gas flowrate (mL/min):  6 (Helium)

Makeup gas flowrate (mL/min):  20 (Helium)

Temperature program:  120°C (3-min hold) to 270°C (10-min hold) at 5°C/min

Injector temperature:  250°C

Detector temperature:  300°C

Injection volume:  2 juL

Solvent:  Hexane

Type of injector:  Flash vaporization

Detector type:  Dual NPD

Range:  1

Attenuation:  64

Type of splitter:   Y-shaped  or Tee

Data system:   Integrator

Hydrogen gas pressure:   20 psi

Bead temperature:  400°C

Bias voltage:   4
                                  8141A -  27                         Revision  1
                                                                September  1994

-------
                               TABLE 10
        QUANTITATION AND CHARACTERISTIC IONS FOR OP PESTICIDES
Compound Name
Quantitation ions
Characteristic ions
Azinphos-methyl
Bolstar (Sulprofos)
Chlorpyrifos
Coumaphos
Demeton-S
Diazinon
Dichlorvos (DDVP)
Dimethoate
Disulfoton
EPN
Ethoprop
Fensulfothion
Fenthion
Malathion
Merphos
Mevinphos
Monocrotophos
Naled
Parathion, ethyl
Parathion, methyl
Phorate
Ronnel
Stirophos
Sulfotepp
TEPP
Tokuthion
160
156
197
109
88
137
109
87
88
157
158
293
278
173
209
127
127
109
291
109
75
285
109
322
99
113
77,132
140,143,113,33
97,199,125.314
97,226,362,21
60,114,170
179,152,93,199,304
79,185,145
93,125,58,143
89,60,61,97,142
169,141,63,185
43,97,41,126
97,125,141,109,308
125,109,93,169
125,127,93,158
57,153,41,298
109,67,192
67,97,192,109
145,147,79
97,109,139,155
125,263,79
121,97,47,260
125,287,79,109
329,331,79
97,65,93,121,202
155,127,81,109
43,162,267,309
                             8141A -  28
                                           Revision  1
                                       September  1994

-------
300.00
250.00
200.00
150.00
100.00
 50.00
  0.00
                  ..



1
e
2

Tetrachlorovinphos
FensuHothion
V... Ji


           . i . . . p . •) i ri | • • i | i . i, i .1 | .1 . , . i . , . ti f i, • , . , ,, , n | , , i ,. , ,, . . . | , . , ,, , ,, , , , , , , , , , , ,, . . . |

            3   5   7  9  11  13  15  17 19  21 23  25  27 29  31  33 35  37 39  41  43 45
 Figure  1.   Chromatogram of  target organophosphorus compounds  from a  15-m DB-21C
 column  with FPD detector.   More compounds  are shown in  Figure 2. See  Table 3 for
 retention times.
                                    8141A - 29
    Revision  1
September 1994

-------
300.00
                                                                    a.
                                                                    UJ
250.00
200.00
150.00
100.00
 50.00
                                   V
  0.00
                                                                 i>
         1   3  5   7  9  11  13  15  17 19 21  23 25 27  29 31 33  35 37  39 41 43 45
  Figure 2.   Chromatogram  of target  organophosphorus  compounds from  a 15-m  DB-210
  column with FPD detector.   More compounds  are shown  in Figure 1.   See Table  3  for
  retention times.
                                     8141A - 30
    Revision  1
September  1994

-------
300.00
250.00
200.00-
150.00-
100.00-
 50.00-
  0.00-J
a.

£
OL


C
 Figure  3.   Chromatogram of  target  organophosphorus  compounds  from  a 15-m  OB-21
 column  with NPD detector.   More compounds  are  shown in Figure  4.  See Table  3  fo
 retention  times.
                                    8141A  -  31
                                                 Revision  1
                                             September  1994

-------
30000-
250.00-
200.00-
150.00-
100.00-
 50.00-
        1  3   5  7  9  11  13  15 17 19  21  23 25  27  29 31  33  35 37 39  41  43 45
  Figure 4.   Chromatogram  of  target organophosphorus compounds  from a  15-m DB-210
  column with NPD detector.   More compounds are  shown in Figure 3.   See  Table 3 for
  retention times.
                                     8141A - 32
    Revision  1
September 1994

-------
      -r
                                                            DB-210
                                n
                                                           DB-5
                                                    •4
Figure 5.  Chromatogram of target organophosphorus compounds on a 30-m DB-5/DB-21C
column pair with NPD detector,  without  Simazine, Atrazine and Carbophenothion.  See
Table 4 for retention times and for GC operating conditions.
                                  8141A - 33
    Revision 1
September 1994

-------
     J
                    is
\J
                                                   DB-210
                                        tl
                                    »
                                    II
                                           n
                                           u
                              «•     a
                                                      41
•i;
M









U 	

II







JL
4







•
il
1




1
17



n


u
IM


II
taJJ
1 "








f
•
.11
1

11
M

»





4t
i.
1






DB-5





4



i
4t
44
JU
Figure 6.  Chromatogram of target organophosphorus compounds on  a 30-m  DB-5/DB-210
column pair with  NPD detector, with Simazine,  Atrazine  and Carbophenothion.   See
Table 4 for retention times and for GC operating conditions.
                                  8141A - 34
                                             Revision  1
                                         September  1994

-------
                                       METHOD  8141A
               ORGANOPHOSPHORUS  COMPOUNDS  BY  GAS  CHROMATOGRAPHY:
                              CAPILLARY  COLUMN  TECHNIQUE
   (    Start     )
7.1.1 Refer to Chapter
  Two for guidance on
choosing the appropriate
  extraction procedure.
    7.1.2 Perform
   solvent exchange
      during K-D
   procedures in all
  extraction methods.
     7.2 Select GC
      conditions.
  7.3 Refer to Method
       8000 for
 calibration techniques.
        I
   7.3.1 Internal or
       external
    calibration may
       be used.
  7.4.1  Add internal
  standard to sample
     if necessary.
     7.4.2 Refer to
   Method 8000, Sec.
   7.6 for instructions
  on analysis sequence,
dilutions, retention times,
    and  identification
        criteria.
  7.4.3 Inject sample.
  7.4.5 Record sample
  volume injected and
  resulting peak sizes.
         I
    7.4.6 Determine
      identity and
    quantity of each
    component peak;
    refer to Method
   8000, Sec. 7.8 for
  calculation equations.
        7.4.7
       Is peak
     detection and
     identification
     prevented by
       interfer-
       ences?
   7.5.1 Perform
appropriate cleanup.
                                  7.5.2 Reanalyze by
                                         GC.
                                         I
                                   (    Stop    J
                                     8141A  - 35
                                                 Revision  1
                                            September  1994

-------
                                 METHOD 8150B

                 CHLORINATED HERBICIDES BY GAS CHROMATOGRAPHY
1.0   SCOPE AND APPLICATION

      1.1   Method 8150  is  a gas chromatographic  (GC)  method  for determining
certain chlorinated acid herbicides.   The  following compounds can be determined
by this method:
      Compound Name                         CAS No.'
      2,4-D                                  94-75-7
      2,4-DB                                 94-82-6
      2,4,5-TP (Silvex)                      93-72-1
      2,4,5-T                                93-76-5
      Dalapon                                75-99-0
      Dicamba                              1918-00-9
      Dichloroprop                          120-36-5
      Dinoseb                                88-85-7
      MCPA                                   94-74-6
      MCPP                                   93-65-2
      a  Chemical  Abstract Services  Registry  Number.

      1.2   Table  1  lists  the method  detection  limit  for  each  compound  in
organic-free reagent  water.  Table 2  lists the estimated quantitation limit (EQL)
for other matrices.

      1.3   When  Method  8150  is used to analyze  unfamiliar  samples,  compound
identifications should  be  supported  by at  least one  additional  qualitative
technique.   This  method describes  analytical conditions  for  a second  gas
chromatographic column that can be  used to confirm  measurements  made  with the
primary column.  Sec. 8.4 provides gas chromatograph/mass spectrometer (GC/MS)
criteria   appropriate   for   the   qualitative   confirmation   of   compound
identifications.

      1.4   Only   experienced   analysts   should  be   allowed  to  work   with
diazomethane due to the  potential hazards associated with its use (the compound
is explosive and carcinogenic).


2.0   SUMMARY OF METHOD

      2.1   Method   8150   provides   extraction,    esterification,  and   gas
chromatographic conditions  for the analysis  of chlorinated  acid  herbicides.
Spiked samples  are used  to  verify the applicability  of  the  chosen extraction
technique to each  new  sample  type.   The esters are hydrolyzed with  potassium


                                  8150B - 1                         Revision 2
                                                                September 1994

-------
hydroxide, and extraneous organic material  is  removed  by a solvent wash.  After
acidification, the acids are extracted with  solvent and converted to their methyl
esters using diazomethane as  the  derivatizing agent.   After excess reagent is
removed, the esters are determined by gas chromatography employing an electron
capture  detector,    microcoulometric detector,  or electrolytic  conductivity
detector  (Goer!Hz  and Lamar,  1967).   The results  are  reported as  the acid
equivalents.

      2.2   The  sensitivity  of Method  8150 usually  depends  on the  level  of
interferences rather than on instrumental limitations.
3.0   INTERFERENCES

      3.1   Refer to Method 8000.

      3.2   Method  interferences  may be  caused  by contaminants  in  solvents,
reagents, glassware, and other sample processing hardware that lead to discrete
artifacts or elevated baselines in gas chromatograms.  All these materials must
be routinely demonstrated to be free  from  interferences under the conditions of
the analysis, by analyzing reagent blanks.

            3.2.1 Glassware must  be  scrupulously cleaned.  Clean each piece of
      glassware  as  soon  as  possible after  use  by rinsing  it  with  the  last
      solvent used in it.  This should be followed  by detergent washing with hot
      water and  rinses  with tap water,  then with  organic-free  reagent water.
      Glassware  should  be  solvent-rinsed  with acetone  and  pesticide-quality
      hexane.  After rinsing and drying, glassware should be sealed and stored
      in  a  clean environment  to  prevent  any  accumulation  of  dust  or  other
      contaminants.    Store  glassware inverted or capped with  aluminum  foil.
      Immediately prior to use, glassware should be rinsed with the next solvent
      to be used.

            3.2.2 The use of high  purity reagents and  solvents helps to minimize
      interference problems.  Purification of solvents by distillation in all-
      glass systems may be required.

      3.3   Matrix  interferences  may  be  caused by  contaminants  that  are
coextracted  from the sample.   The  extent  of  matrix interferences  will  vary
considerably from waste to waste,  depending upon the nature  and diversity of the
waste being sampled.

      3.4   Organic acids, especially chlorinated acids, cause the most direct
interference with the determination.  Phenols, including chlorophenols, may also
interfere with this procedure.

      3.5   Alkaline hydrolysis and subsequent  extraction of  the basic solution
remove many chlorinated hydrocarbons and phthalate esters that might otherwise
interfere with the electron capture  analysis.

      3.6   The  herbicides, being  strong organic acids,  react  readily  with
alkaline substances and may be lost during analysis.   Therefore, glassware and
glass wool  must  be  acid rinsed,  and sodium  sulfate  must  be  acidified  with
sulfuric acid prior to use to avoid  this possibility.

                                   8150B - 2                         Revision 2
                                                                September 1994

-------
      3.7   Sample extracts  should  be dry  prior  to methylation or  else poor
recoveries will be obtained.
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  chromatograph,  analytical  system  complete  with  gas
      chromatograph  suitable  for   on-column   injections   and  all   required
      accessories, including detectors, analytical columns,  recorder, gases, and
      syringes.  A data system for measuring peak heights and/or peak areas is
      recommended.

            4.1.2 Columns

                  4.1.2.1     Column la and Ib - 1.8 m x 4 mm ID glass, packed
            with  1.5%  SP-2250/1.95% SP-2401 on Supelcoport  (100/120  mesh)  or
            equivalent.

                  4.1.2.2     Column 2 - 1.8 m x 4 mm ID glass, packed with 5%
            OV-210 on Gas Chrom Q (100/120 mesh) or equivalent.

                  4.1.2.3     Column 3  -  1.98  m x 2 mm  ID glass,  packed with
            0.1% SP-1000 on 80/100 mesh Carbopack C or equivalent.

            4.1.3 Detector  - Electron capture (BCD).

      4.2   Erlenmeyer flasks - 250  and 500 ml  Pyrex,  with 24/40 ground glass
joint.

      4.3   Beaker - 500 mL.

      4.4   Diazomethane generator -  Refer to Sec. 7.4 to  determine which method
of diazomethane generation should be used  for a particular application.

            4.4.1 Diazald kit - recommended for the generation of diazomethane
      using the procedure given in Sec. 7.4.2  (Aldrich Chemical  Co.,  Cat.  No.
      210,025-2 or equivalent).

            4.4.2 Assemble from two 20  x 150 mm test tubes, two Neoprene rubber
      stoppers, and a  source of  nitrogen.   Use  Neoprene rubber  stoppers with
      holes drilled in  them  to accommodate glass delivery tubes.  The exit tube
      must  be  drawn to  a  point  to bubble diazomethane  through the  sample
      extract.   The generator assembly  is shown  in Figure 1.  The procedure for
      use of this type  of generator is  given in Sec.  7.4.3.

      4.5   Vials - 10 to  15 mL,  amber glass,  with Teflon  lined  screw cap or
crimp top.

      4.6   Separatory funnel - 2000 mL,  125 mL, and 60 mL.

      4.7   Drying column -  400 mm x  20 mm ID Pyrex chromatographic column with
Pyrex glass wool at bottom and a  Teflon stopcock.

                                  8150B - 3                         Revision 2
                                                                September 1994

-------
      NOTE: Fritted  glass  discs are  difficult  to decontaminate  after highly
            contaminated extracts  have  been passed through.   Columns without
            frits may  be  purchased.   Use  a  small  pad of Pyrex glass wool  to
            retain the  adsorbent.   Prewash the  glass  wool  pad with  50  ml  of
            acetone  followed by 50 ml of  elution solvent  prior to packing the
            column with adsorbent.

      4.8   Kuderna-Danish (K-D) apparatus

            4.8.1 Concentrator tube - 10 ml, graduated  (Kontes K-570050-1025 or
      equivalent).   A  ground  glass stopper is used to prevent evaporation  of
      extracts

            4.8.2 Evaporation   flask  -   500   ml  (Kontes   K-570001-500  or
      equivalent).   Attach  to  concentrator  tube with  springs,  clamps  or
      equivalent.

            4.8.3 Snyder column  - Three  ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.8.4 Snyder  column  -  Two ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.8.5 Springs  -  1/2 inch (Kontes K-662750 or equivalent).

      4.9   Boiling chips  - Solvent extracted, approximately  10/40 mesh (silicon
carbide or equivalent).

      4.10  Water  bath  -  Heated,  with  concentric  ring  cover,  capable  of
temperature control   (+ 5°C).   The  bath should  be  used  in a hood.

      4.11  Microsyringe - 10 /xL.

      4.12  Wrist shaker - Burrell Model 75 or equivalent.

      4.13  Glass wool - Pyrex, acid washed.

      4.14  Balance  - Analytical, capable of accurately weighing to 0.0001 g.

      4.15  Syringe  - 5 ml.

      4.16  Glass rod.
5.0   REAGENTS

      5.1   Reagent  grade  inorganic  chemicals  shall  be  used  in  all  tests.
Unless otherwise indicated,  it is intended that  all reagents shall  conform to the
specifications of the Committee on  Analytical Reagents of the American Chemical
Society, where  such  specifications are available.  Other  grades  may be used,
provided it is first  ascertained that the reagent is  of sufficiently  high purity
to permit its use without lessening the accuracy of the determination.
8150B - 4
                                                                    Revision 2
                                                                September 1994

-------
      5.2   Organic-free reagent water.  All references to water in  this  method
refer to organic-free reagent water, as defined  in Chapter One.

      5.3   Sulfuric acid  solution

            5.3.1  ((1:1) (v/v)) - Slowly add 50 ml H2S04 (sp.  gr. 1.84)  to 50  ml
      of organic-free reagent water.

            5.3.2  ((1:3) (v/v)) - Slowly add 25 ml H2S04 (sp.  gr. 1.84)  to 75  ml
      of organic-free reagent water.

      5.4   Hydrochloric   acid   ((1:9)   (v/v)),   HC1.     Add   one   volume   of
concentrated HC1 to 9 volumes of organic-free  reagent  water.

      5.5   Potassium  hydroxide  solution (KOH)  - 37%  aqueous  solution  (w/v).
Dissolve 37 g  potassium hydroxide pellets  in  organic-free reagent  water, and
dilute to 100 ml.

      5.6   Carbitol (Diethylene glycol  monoethyl ether), C2H5OCH2CH2OCH2CH2OH.
 Available from Aldrich Chemical Co.

      5.7   Solvents

            5.7.1  Acetone,  CH3COCH3 - Pesticide quality or equivalent.

            5.7.2  Methanol,  CH3OH - Pesticide  quality  or  equivalent.

            5.7.3  Isooctane,   (CH3)3CCH2CH(CH3)2   -    Pesticide   quality   or
      equivalent.

            5.7.4  Hexane,  C6HU  -  Pesticide  quality or  equivalent.

            5.7.5  Diethyl  Ether,  C2H5OC2H5.    Pesticide quality or  equivalent.
      Must  be  free of  peroxides as  indicated  by test  strips  (EM Quant,  or
      equivalent).  Procedures for  removal  of peroxides  are provided  with the
      test strips.  After cleanup, 20 ml of ethyl alcohol preservative must  be
      added to each liter of ether.

      5.8   Sodium sulfate  (granular,  acidified,  anhydrous),  Na2S04. Purify  by
heating at  400°C for 4  hours in  a shallow  tray,  or  by precleaning  the  sodium
sulfate with  methylene chloride.   If the  sodium sulfate  is  precleaned with
methylene chloride, a method blank must be analyzed, demonstrating that  there  is
no interference  from  the sodium sulfate.   Acidify  by  slurrying  100  g  sodium
sulfate with enough diethyl  ether to just cover the  solid; then  add 0.1  ml  of
concentrated sulfuric acid  and mix thoroughly.  Remove the ether under a vacuum.
Mix 1  g  of the  resulting  solid  with  5  ml  of organic-free  reagent water and
measure the pH of the mixture.  It must be below a pH  of  4.  Store  at  130°C.

      5.9   N-Methyl-N-nitroso-p-toluenesulfonamide (Diazald), CH3C6H4S02N(CH3)NO.
 Available from Aldrich Chemical  Co.

      5.10  Silicic acid.   Chromatographic  grade, nominal 100 mesh.   Store  at
130°C.

                                  8150B  - 5                         Revision 2
                                                                September 1994

-------
      5.11  Stock standard solutions -  Stock standard solutions can be prepared
from pure standard materials or purchased as  certified  solutions.

            5.11.1      Prepare stock standard solutions by accurately weighing
      about 0.0100 g of pure  acids.  Dissolve the acids  in  pesticide quality
      acetone and dissolve the esters in  10% acetone/isooctane (v/v) and dilute
      to volume in a  10 ml volumetric flask.   Larger volumes can be used at the
      convenience of the  analyst.   If  compound purity is certified  at  96% or
      greater,  the  weight  can  be  used  without correction  to calculate  the
      concentration  of the  stock  standard.    Commercially  prepared  stock
      standards can be used at any  concentration if they  are  certified  by the
      manufacturer or by an independent source.

            5.11.2      Transfer the stock standard solutions  into vials with
      Teflon  lined screw  caps or crimp  tops.   Store  at  4°C and  protect from
      light.  Stock standard solutions should  be checked frequently  for signs of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.11.3      Stock standard solutions of the derivatized acids must
      be replaced after 1 year, or  sooner, if comparison  with  check standards
      indicates a problem.  Stock standard solutions  of the free acids degrade
      more  quickly  and should  be  replaced   after two  months,  or  sooner  if
      comparison with check standards indicates a problem.

      5.12  Calibration standards -  A minimum of five calibration standards  for
each parameter  of interest should  be  prepared through  dilution of  the stock
standards  with  diethyl ether.    One  of the  concentrations  should  be  at  a
concentration near,  but  above,  the method  detection  limit.   The  remaining
concentrations should correspond to the expected range  of concentrations found
in real  samples or should  define  the  working  range of  the GC.   Calibration
solutions must be replaced after  six months, or sooner if comparison with check
standards indicates a problem.

      5.13  Internal  standards (if  internal  standard  calibration is used) - To
use this approach, the analyst must select one or more  internal  standards that
are similar in  analytical   behavior  to  the compounds of  interest.   The analyst
must further  demonstrate  that  the measurement of the  internal  standard is not
affected by method or matrix  interferences.   Because of  these  limitations, no
internal standard can be suggested that is applicable  to all  samples.

            5.13.1      Prepare  calibration   standards  at a  minimum of five
      concentrations for each parameter of interest as  described in Sec. 5.12.

            5.13.2      To  each  calibration  standard,  add  a  known  constant
      amount  of one or more  internal  standards,  and dilute  to  volume with
      hexane.

            5.13.3      Analyze each calibration standard per Sec. 7.0.

      5.14  Surrogate standards - The analyst  should monitor the performance of
the extraction,  cleanup (when used),  and analytical system,  and the effectiveness
of  the  method  in dealing  with  each  sample  matrix  by  spiking  each sample,


                                   8150B  - 6                         Revision 2
                                                                September 1994

-------
standard,  and organic-free  reagent water  blank  with  one  or  two  herbicide
surrogates (e.g.  herbicides that are not  expected  to be present in the sample).
The surrogates selected should elute over the range of the temperature program
used in this method.  2,4-Dichlorophenylacetic acid (DCAA) is recommended as a
surrogate  compound.   Deuterated  analogs of  analytes should  not be  used  as
surrogates for gas chromatographic analysis due to coelution problems.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See  the  introductory  material  to this Chapter,  Organic  Analytes,
Sec. 4.1.  Extracts  must  be  stored  under refrigeration  and analyzed  within 40
days of extraction.


7.0  PROCEDURE

      7.1   Preparation of waste samples

            7.1.1  Extraction

                   7.1.1.1     Follow Method 3580 except use diethyl ether as the
            dilution solvent, acidified anhydrous  sodium sulfate,  and acidified
            glass wool.

                   7.1.1.2     Transfer 1.0 mL (a lesser volume or  a dilution may
            be required if herbicide concentrations are high) to a 250 mL ground
            glass-stoppered Erlenmeyer  flask.  Proceed to Sec. 7.2.2 hydrolysis.

      7.2   Preparation of soil, sediment, and other solid samples

            7.2.1  Extraction

                   7.2.1.1     To a  500 mL,  wide mouth Erlenmeyer flask add 50
            g (dry weight as  determined in Method  3540, Sec. 7.2.1) of the well
            mixed, moist  solid  sample.   Adjust  the pH to  2  (See Method  9045)
            with  concentrated HC1  and  monitor  the  pH  for  15   minutes  with
            occasional stirring.  If necessary, add additional HC1 until the pH
            remains at 2.

                   7.2.1.2     Add  20  mL acetone  to  the  flask  and  mix  the
            contents with the wrist shaker  for 20 minutes.   Add  80 mL diethyl
            ether to the same flask  and shake  again for 20 minutes.  Decant the
            extract and measure the volume of solvent recovered.

                  7.2.1.3     Extract  the  sample  twice  more  using  20 mL  of
            acetone followed  by 80 mL of  diethyl ether.  After addition of each
            solvent, the  mixture  should  be  shaken with  the wrist shaker  for
            10 minutes and the acetone-ether extract  decanted.

                  7.2.1.4     After the third extraction, the volume of extract
            recovered should  be at  least 75% of the volume of added  solvent.
            If this is not the  case, additional extractions  may  be necessary.
            Combine  the  extracts  in a 2 liter  separatory funnel containing

                                  8150B  - 7                        Revision 2
                                                                September  1994

-------
250 ml of reagent water.   If an  emulsion  forms,  slowly  add  5 g of
acidified sodium sulfate (anhydrous) until  the solvent-water mixture
separates.   A  quantity of  acidified  sodium sulfate equal  to the
weight of the sample may be added, if necessary.

      7.2.1.5     Check the pH of the extract.   If it is not at or
below pH  2, add more concentrated  HC1   until  stabilized  at  the
desired pH.  Gently mix the contents  of the separatory  funnel for
1 minute  and allow  the layers to  separate.  Collect the aqueous
phase in  a clean beaker  and  the extract  phase  (top layer)  in  a
500 ml ground glass-stoppered  Erlenmeyer flask.   Place the aqueous
phase back into the  separatory  funnel  and  re-extract using 25 ml of
diethyl  ether.   Allow the layers to separate and discard the aqueous
layer.  Combine the ether extracts in the  500 ml  Erlenmeyer flask.

      7.2.1.6     An   alternative   extraction   procedure   using
ultrasonic extraction can be found in Sec. 7.2 of Method 8151.

7.2.2 Hydrolysis

      7..2.2.1     Add 30 ml of organic-free reagent water, 5 ml of
37% KOH, and one or two clean  boiling chips to the flask.   Place a
three ball Snyder column on the flask, evaporate  the diethyl ether
on a water bath, and continue  to heat until the hydrolysis step is
completed (usually 1 to 2 hours).

      7.2.2.2     Remove the flask from the water bath and allow to
cool.  Transfer the water solution to a 125 ml separatory funnel and
extract the  basic  solutions once with  40 mL and  then  twice with
20 ml of  diethyl ether.   Allow sufficient  time for  the  layers to
separate and discard the ether layer  each time.   The phenoxy-acid
herbicides remain soluble in the aqueous phase as potassium salts.

7.2.3 Solvent cleanup

      7.2.3.1     Adjust  the  pH  to  2 by  adding  5  ml  cold  (4°C)
sulfuric acid (1:3)  to the  separatory  funnel.  Be sure to check the
pH at this point.  Extract  the  herbicides  once with 40 ml and twice
with 20 ml of diethyl ether.  Discard the  aqueous phase.

      7.2.3.2     Combine  ether  extracts   in  a 125  ml  Erlenmeyer
flask  containing 5-7  g of acidified  anhydrous   sodium  sulfate.
Stopper  and allow  the  extract   to   remain  in  contact  with  the
acidified sodium sulfate.  If concentration and esterification are
not to be performed immediately,  store the sample overnight in the
refrigerator.

      NOJE: The  drying step is very  critical  to  ensuring complete
            esterification.  Any moisture  remaining  in  the ether
            will result in low herbicide recoveries. The amount of
            sodium   sulfate  is   adequate  if   some  free=flowing
            crystals are visible  when swirling the flask.  If all
            the  sodium sulfate  solidifies in a  cake,  add  a few
                       8150B  -  8                         Revision 2
                                                    September 1994

-------
                  additional grams of acidified sodium sulfate and again
                  test by swirling.  The 2  hour drying time is a minimum,
                  however, the extracts may be held overnight in contact
                  with the sodium  sulfate.

            7.2.3.3     Transfer  the  ether  extract,  through  a  funnel
      plugged  with  acid  washed  glass  wool,  into  a 500  ml K-D  flask
      equipped with a 10 ml concentrator tube.  Use a glass rod to crush
      caked sodium  sulfate during the  transfer.   Rinse  the Erlenmeyer
      flask and  column  with 20-30 mL  of diethyl  ether to  complete the
      quantitative transfer.

            7.2.3.4     Add one or two  clean  boiling chips  to the flask
      and attach a three ball  Snyder  column.   Prewet the Snyder column by
      adding about 1 ml  of diethyl  ether to the top.  Place the apparatus
      on  a  hot water bath  (60°-65°C)  so  that the concentrator  tube is
      partially  immersed  in  the  hot  water and the  entire  lower rounded
      surface  of the flask  is bathed  in  vapor.   Adjust the  vertical
      position of the apparatus and  the water temperature,  as required,
      to complete the concentration in  15-20 minutes.  At the proper rate
      of distillation the balls of the column will actively chatter, but
      the chambers  will  not  flood.  When  the apparent  volume of liquid
      reaches 1 mL,  remove the  K-D  apparatus from the water bath and allow
      it to drain and cool for at  least 10 minutes.

            7.2.3.5     Remove the Snyder  column and rinse the flask and
      its lower joints into the concentrator tube with 1-2 ml of diethyl
      ether.   A  5 ml  syringe is  recommended  for  this operation.   Add a
      fresh boiling  chip,  attach a micro Snyder column to the concentrator
      tube, and prewet the column by adding 0.5 ml of ethyl  ether to the
      top.  Place the micro K-D apparatus  on  the  water  bath  so that the
      concentrator tube is partially immersed in  the hot  water.  Adjust
      the vertical  position of the  apparatus and the water temperature as
      required  to  complete  concentration  in  5-10 minutes.    When  the
      apparent volume of the liquid reaches 0.5 ml, remove the micro K-D
      from the bath and  allow it  to  drain and cool.   Remove the Snyder
      column  and add  0.1  ml  of  methanol.    Rinse  the walls of  the
      concentrator tube  while adjusting  the extract volume to 1.0 ml with
      diethyl  ether.  Proceed to Sec. 7.4  for esterification.

7.3   Preparation of aqueous samples

      7.3.1 Extraction

            7.3.1.1      Using  a  1 liter  graduated cylinder, measure 1
      liter (nominal)  of sample,  record the sample volume to the nearest
      5 ml, and  quantitatively transfer it to the separatory funnel. If
      high concentrations are  anticipated, a  smaller volume  may be used
      and then diluted with organic-free reagent  water to 1 liter.  Adjust
      the pH to less than 2 with  sulfuric acid (1:1).

            7.3.1.2     Add 150 ml of diethyl  ether to the sample bottle,
      seal, and  shake for  30  seconds to rinse the walls.  Transfer the
                            8150B  - 9                         Revision 2
                                                          September 1994

-------
solvent wash  to  the separatory  funnel  and extract the  sample  by
shaking the funnel  for 2 minutes with  periodic  venting to release
excess pressure.   Allow the organic layer to separate from the water
layer  for  a  minimum  of  10  minutes.    If  the  emulsion  interface
between layers is more than one third the size of the solvent layer,
the analyst must  employ mechanical techniques to complete the phase
separation.  The optimum  technique depends  upon  the sample and may
include stirring,  filtration  of the emulsion through  glass wool,
centrifugation, or other  physical methods.  Drain the aqueous phase
into a 1 liter Erlenmeyer flask.  Collect the solvent  extract in a
250 ml ground glass Erlenmeyer flask containing 2 ml  of 37% KOH.
Approximately 80 mL of the diethyl  ether will remain  dissolved  in
the aqueous phase.

      7.3.1.3      Repeat the extraction two more times using 50 ml
of diethyl  ether  each  time.  Combine the extracts in the Erlenmeyer
flask.  (Rinse the 1  liter flask with  each  additional  aliquot  of
extracting solvent.)

7.3.2 Hydrolysis

      7.3.2.1      Add one or  two clean  boiling  chips  and 15 ml  of
organic-free  reagent water to  the 250  ml flask  and attach a three
ball Snyder column.  Prewet the Snyder column by adding about 1  mL
of diethyl  ether to the top of the column.   Place the  apparatus  on
a hot water bath  (60°-65°C) so that the bottom of  the flask is bathed
with hot water vapor.  Although the diethyl ether will  evaporate in
about  15 minutes,  continue heating  until  the  hydrolysis  step  is
completed  (usually  1  to  2  hours).    Remove  the apparatus  and let
stand at room temperature for at least 10 minutes.

      7.3.2.2      Transfer the solution to a 60 ml separatory funnel
using  5-10 ml  of  organic-free  reagent water.    Wash  the  basic
solution twice  by  shaking  for  1 minute with  20  ml   portions  of
diethyl ether.   Discard  the organic  phase.   The herbicides remain
in the aqueous phase.

7.3.3 Solvent cleanup

      7.3.3.1      Acidify the  contents of the separatory funnel  to
pH 2 by adding 2  ml of cold (4°C) sulfuric acid (1:3).  Test with pH
indicator paper.   Add 20  ml diethyl  ether and shake vigorously for
2 minutes.   Drain the  aqueous  layer  into  a  250 mL Erlenmeyer flask,
and pour the organic layer into a 125 mL  Erlenmeyer  flask containing
about  5-7  g of  acidified sodium sulfate.    Repeat the extraction
twice  more with   10 mL  aliquots of  diethyl  ether, combining all
solvent in  the 125 mL flask.  Allow the extract to remain  in contact
with the sodium  sulfate for approximately 2 hours.

      NOTE: The  drying step  is very critical  to ensuring complete
            esterification.   Any moisture remaining  in  the ether
            will result in  low herbicide recoveries.  The amount of
            sodium sulfate  is  adequate  if  some  free  flowing
                      8150B - 10                        Revision 2
                                                    September 1994

-------
                   crystals  are  visible  when swirling the flask.  If all
                   the  sodium sulfate  solidifies  in  a  cake,  add  a few
                   additional grams of acidified sodium sulfate  and again
                   test by swirling.  The 2 hour drying time  is  a minimum,
                   however,  the  extracts may be  held  overnight  in contact
                   with the  sodium  sulfate.

             7.3.3.2      Transfer  the  ether extract,  through  a  funnel
      plugged  with  acid washed glass  wool,   into   a 500  ml  K-D  flask
      equipped with  a 10 ml concentrator  tube.  Use  a glass  rod to crush
      caked  sodium sulfate during the  transfer.    Rinse the  Erlenmeyer
      flask  and  column   with 20-30 ml  of diethyl  ether  to  complete the
      quantitative transfer.

             7.3.3.3      Add one or two  clean  boiling chips to the flask
      and attach a three ball Snyder  column.  Prewet  the  Snyder column by
      adding about 1 mL of diethyl  ether to the top.   Place the  apparatus
      on  a  hot water bath  (60°-65°C)  so  that  the  concentrator tube is
      partially  immersed in the hot  water  and  the  entire  lower rounded
      surface  of the flask is bathed  in  vapor.    Adjust  the vertical
      position of  the apparatus and  the water temperature,  as required,
      to complete the concentration in  15-20 minutes. At the proper rate
      of distillation the balls of the column will actively chatter, but
      the chambers  will  not flood.  When the  apparent  volume of liquid
      reaches 1 ml, remove the K-D apparatus from the water bath and allow
      it to drain and cool  for  at least 10 minutes.

             7.3.3.4      Remove  the Snyder column  and rinse the  flask and
      its lower joints into the concentrator tube with 1-2 ml  of diethyl
      ether.   A  5  ml syringe is recommended for  this operation.   Add a
      fresh boiling chip, attach a micro Snyder column to  the concentrator
      tube, and prewet the  column by adding 0.5 mL of ethyl ether to the
      top.   Place  the micro K-D apparatus on  the  water  bath so that the
      concentrator tube  is  partially immersed  in  the hot water.  Adjust
      the vertical  position of the  apparatus and the  water temperature as
      required  to  complete  concentration  in  5-10  minutes.    When  the
      apparent volume of the liquid reaches 0.5 ml,  remove the micro K-D
      from the bath  and  allow  it  to drain  and  cool.  Remove the Snyder
      column  and  add  0.1  ml   of  methanol.    Rinse the  walls of  the
      concentrator tube while adjusting  the extract  volume to 1.0 mL with
      diethyl ether.

7.4   Esterification

      7.4.1 Two methods  may be  used  for the generation  of diazomethane:
the bubbler method  (set  up shown in Figure 1) and  the Diazald kit method.
The bubbler  method is suggested when  small batches  (10-15)  of samples
require esterification.   The bubbler  method works  well with samples that
have low concentrations  of herbicides (e.g.  aqueous  samples)  and is safer
to use than the Diazald kit procedure.  The Diazald kit method is good for
large quantities  of samples  needing  esterification.   The  Diazald  kit
method is more  effective  than the bubbler method for  soils or  samples that
may contain high concentrations  of herbicides (e.g.,  samples  such as soils
                            8150B - 11                        Revision 2
                                                          September 1994

-------
that result -in yellow  extracts  following  hydrolysis  may be difficult to
handle by the  bubbler method).  The diazomethane derivatization (U.S. EPA,
1971) procedures, described below,  will  react efficiently with all of the
chlorinated herbicides  described in this method and should be used only by
experienced analysts,  due  to the potential hazards  associated  with its
use.  The following precautions should be taken:

      CAUTION:    Diazomethane  is a  carcinogen   and  can   explode  under
                  certain  conditions.

                  Use  a safety  screen.
                  Use  mechanical pipetting aides.
                  Do not heat above 90°C -- EXPLOSION may  result.
                  Avoid grinding surfaces, ground glass joints,  sleeve
                  bearings,  glass stirrers --     EXPLOSION may result.
                  Store away from alkali metals -- EXPLOSION may result.
                  Solutions  of  diazomethane  decompose rapidly in  the
                  presence  of  solid  materials  such  as copper  powder,
                  calcium chloride, and boiling chips.

      7.4.2 Diazald kit method  - Instructions for preparing diazomethane
are provided with the generator kit.

            7.4.2.1     Add  2 mL of diazomethane  solution and let sample
      stand for  10 minutes with occasional swirling.

            7.4.2.2     Rinse  inside wall  of the  ampule with  several
      hundred  ^L   of   diethyl   ether.     Allow   solvent   to  evaporate
      spontaneously at room  temperature to about  2 mL.

            7.4.2.3     Dissolve the  residue in 5 mL of hexane.  Analyze
      by gas chromatography.

      7.4.3 Bubbler  method  -   Assemble  the  diazomethane  bubbler  (see
Figure 1).

            7.4.3.1     Add  5 mL of diethyl ether  to  the first test tube.
      Add 1 mL of diethyl  ether, 1 mL of carbitol,  1.5 mL of 37% KOH, and
      0.1-0.2 g  Diazald to  the  second test tube.   Immediately place the
      exit tube into the concentrator  tube containing the sample extract.

            Apply nitrogen flow (10 mL/min) to bubble diazomethane through
      the extract for 10 minutes or until the yellow color of diazomethane
      persists.     The  amount  of   Diazald  used   is   sufficient  for
      esterification   of   approximately   three   sample   extracts.     An
      additional  0.1-0.2  g  of  Diazald  may be added  (after  the initial
      Diazald is consumed) to extend  the generation of the diazomethane.
      There is sufficient KOH present in the original solution to perform
      a maximum  of  approximately 20 minutes of total esterification.

            7.4.3.2     Remove  the concentrator  tube and  seal  it with  a
      Neoprene or Teflon stopper. Store at room temperature in  a hood for
      20 minutes.


                             8150B - 12                        Revision  2
                                                          September 1994

-------
                  7.4.3.3      Destroy  any  unreacted  diazomethane  by  adding
            0.1-0.2 g  silicic  acid  to  the concentrator tube.   Allow to stand
            until the evolution of nitrogen gas has  stopped.  Adjust  the sample
            volume to  10.0  ml  with  hexane.   Stopper the concentrator tube and
            store  refrigerated if  further processing  will  not  be  performed
            immediately.   It  is  recommended  that the  methylated  extracts be
            analyzed immediately to minimize the trans-esterification and other
            potential reactions that may occur. Analyze by gas chromatography.

      7.5   Gas chromatographic conditions  (Recommended)

            7.5.1 Column la

            Carrier gas (5% methane/95% argon)  flow rate:    70 mL/min
            Temperature program:     185°C, isothermal.

            7.5.2 Column Ib

            Carrier gas (5% methane/95% argon)  flow rate:    70 mL/min
            Initial temperature:     140°C, hold for 6 minutes
            Temperature program:     140°C to 200°C at  10°C/min, hold  until last
                                     compound  has  eluted.

            7.5.3 Column 2

            Carrier gas (5% methane/95% argon)  flow rate:    70 mL/min
            Temperature program:     185°C, isothermal.

            7.5.4 Column 3

            Carrier gas (ultra-high  purity N2) flow rate:     25 mL/min
            Initial temperature:     100°C, no hold
            Temperature program:     100°C to 150°C at  10°C/min, hold  until last
                                     compound  has  eluted.

      7.6   Calibration  -   Refer   to   Method  8000  for  proper  calibration
techniques.  Use Table  1 and especially  Table 2 for guidance on selecting the
lowest point on the calibration curve.

            7.6.1 The  procedure  for internal  or  external calibration  may be
      used.  Refer  to Method 8000 for a  description of  each of these procedures.

            7.6.2 The following gas chromatographic  columns are recommended for
      the compounds indicated:


            Analyte            Column           Analyte           Column

            Dicamba            la,2             Dalapon             3
            2,4-D              la,2             MCPP                 Ib
            2,4,5-TP           la,2             MCPA                 Ib
            2,4,5-T            la,2             Dichloroprop         Ib
            2,4-DB             la               Dinoseb             Ib


                                  8150B - 13                        Revision 2
                                                                September 1994

-------
      7.7   Gas chromatographic analysis

            7.7.1 Refer to Method 8000.   If  the  internal  standard calibration
      technique is used,  add 10 /xL of internal standard to the sample prior to
      injection.

            7.7.2 Method 8000  provides  instructions on  the analysis sequence,
      appropriate dilutions,  establishing daily  retention time  windows,  and
      identification criteria.  Include a mid-concentration  check standard after
      each group of 10 samples in the analysis sequence.

            7.7.3 Examples  of chromatograms  for  various  chlorophenoxy  acid
      herbicides are shown in Figures 2 through 4.

            7.7.4 Record the sample volume injected and the  resulting peak sizes
      (in area units or peak heights).

            7.7.5 Using either the internal  or external  calibration procedure
      (Method 8000), determine the identity and quantity of each component peak
      in the  sample  chromatogram  which corresponds to the compounds  used  for
      calibration purposes.

            7.7.6 If calibration standards have been analyzed  in the same manner
      as the samples (e.g. have undergone hydrolysis and esterification), then
      the calculation  of  concentration given  in  Method 8000 should  be  used.
      However,  if calibration  is  done using  standards made  from  methyl  ester
      compounds (compounds not esterified by  application of this method), then
      the  calculation  of  concentration  must include  a  correction  for  the
      molecular weight of the methyl  ester versus the  acid herbicide.

            7.7.7 If peak  detection   and  identification  are  prevented due to
      interferences, further  cleanup is  required.   Before using  any cleanup
      procedure, the  analyst  must process a  series of standards  through  the
      procedure to validate elution  patterns  and  the  absence of interferences
      from reagents.


8.0   QUALITY CONTROL

      8.1   Refer  to  Chapter  One for specific  quality   control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was performed,  follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Procedures to  check the   GC  system operation  are found  in  Method
8000.

            8.2.1 Select a representative spike concentration for each compound
      (acid or ester) to  be measured.   Using stock standards, prepare a quality
      control check sample concentrate in  acetone  1,000 times more concentrated
      than the selected concentrations.
                                  8150B - 14                        Revision 2
                                                                September 1994

-------
            8.2.2 Table 3  indicates single operator accuracy and precision for
      this  method.   Compare  the results obtained  with  the results  given in
      Table 3 to determine if the data quality is acceptable.

      8.3   Calculate  surrogate  standard  recovery  on  all  samples,  blanks, and
spikes.   Determine if  the recovery is  within limits (limits  established by
performing QC procedures outlined in Method 8000).

            8.3.1 If recovery is not within limits,  the following procedures are
      required.

                  •     Check to be  sure  there are no errors in calculations,
                        surrogate  solutions  and  internal  standards.   Also,
                        check instrument  performance.

                  •     Recalculate  the  data and/or  reanalyze the  extract if
                        none of  the above checks reveal a problem.

                  •     Re-extract  and re-analyze  the  sample  if none  of the
                        above  are a problem or flag the data  as  "estimated
                        concentration".

      8.4   GC/MS confirmation

            8.4.1 GC/MS techniques  should be  judiciously employed  to support
      qualitative identifications made with this method.   Refer to Method 8270
      for the appropriate GC/MS operating conditions  and analysis procedures.

            8.4.2 When  available,  chemical   ionization  mass  spectra may  be
      employed to aid the qualitative identification  process.

            8.4.3 Should  these  MS  procedures  fail  to  provide  satisfactory
      results, additional  steps may be taken before  reanalysis.  These  steps may
      include the use  of alternate packed  or  capillary GC columns or additional
      cleanup.


9.0   METHOD PERFORMANCE

      9.1   In  a  single   laboratory,  using  organic-free  reagent  water  and
effluents from publicly owned  treatment  works (POTW), the  average  recoveries
presented in  Table  3  were obtained.   The standard deviations of  the percent
recoveries of these measurements are also included  in Table  3.
10.0  REFERENCES

1.    U.S. EPA, National  Pollutant Discharge Elimination  System,  Appendix A,
      Fed. Reg.,  38,  No.  75,  Pt. II,  Method  for  Chlorinated Phenoxy  Acid
      Herbicides in Industrial Effluents, Cincinnati, Ohio, 1971.

2.    Goerlitz, D.G., and W.L.  Lamar,  "Determination of Phenoxy Acid Herbicides
      in Water  by  Electron Capture and  Microcoulometric  Gas  Chromatography,"
      U.S. Geol. Survey Water Supply Paper,  1817-C,  1967.

                                  8150B - 15                        Revision 2
                                                                September 1994

-------
Burke, J.A.,  "Gas  Chromatography for Pesticide Residue  Analysis;   Some
Practical Aspects,"   Journal  of the Association  of Official  Analytical
Chemists, 48, 1037, 1965.

U.S.  EPA,  "Extraction and  Cleanup Procedure  for the  Determination  of
Phenoxy Acid Herbicides  in  Sediment,"  EPA  Toxicant and Analysis Center,
Bay St. Louis, Mississippi, 1972.

"Pesticide Methods Evaluation,"  Letter Report #33 for EPA Contract No. 68-
03-2697.    Available   from   U.S.   Environmental   Protection   Agency,
Environmental Monitoring and Support Laboratory, Cincinnati, Ohio 45268.

Eichelberger, J.W., L.E. Harris, and W.L.  Budde,  "Reference Compound to
Calibrate   Ion   Abundance   Measurement   in    Gas   Chromatography-Mass
Spectrometry," Analytical Chemistry, 4_7, 995,  1975.

Glaser,  J.A.  et.al.,  "Trace Analysis  for Wastewaters,"  Environmental
Science & Technology, 15, 1426,  1981.

Gurka, D.F, Shore,  F.L., Pan, S-T,  "Single Laboratory Validation of EPA
Method 8150  for  Determination  of  Chlorinated  Herbicides  in  Hazardous
Waste", JAOAC., 69, 970, 1986.

U.S.  EPA,  "Method 615.  The Determination  of Chlorinated  Herbicides  in
Industrial and Municipal Wastewater," Environmental Monitoring and Support
Laboratory, Cincinnati, Ohio, 45268, June  1982.
                            8150B  -  16                        Revision 2
                                                          September  1994

-------
                                   TABLE 1.
                CHROMATOGRAPHIC CONDITIONS  AND DETECTION LIMITS
                           FOR CHLORINATED  HERBICIDES
Compound
	Retention time (min)"

Col.la   Col.Ib  Col.2   Col.3
"Column conditions are given in Sees.  4.1 and 7.5.
                                   TABLE 2.
                    DETERMINATION OF  ESTIMATED QUANTITATION
                      LIMITS  (EQL) FOR VARIOUS MATRICES3
  Method
 detection
limit (//g/L)
2,4-D
2,4-DB
2,4,5-T
2,4,5-TP (Silvex)
Dalapon
Dicamba
Dichloroprop
Dinoseb
MCPA
MCPP
2.0
4.1
3.4
2.7
-
1.2
-
-
-
-

-
-
-
-
-
4.8
11.2
4.1
3.4
1.6
-
2.4
2.0
5.0
1.0
-
-
-
-
1.2
0.91
0.20
0.17
5.8
0.27
0.65
0.07
249
192
    Matrix
                                Factor
Ground water (based on one liter sample size)
Soil/sediment and other solids
Waste samples
                                   10
                                  200
                              100,000
aEQL = [Method detection limit  (see Table  1)]  X  [Factor found in this table].
For non-aqueous samples, the factor is on a wet weight basis.  Sample EQLs are
highly matrix  dependent.   The  EQLs to  be determined herein  are provided for
guidance and may not always be achievable.
                                  8150B - 17
                                      Revision 2
                                  September 1994

-------
                                   TABLE 3.
                    SINGLE OPERATOR ACCURACY AND PRECISION8
Compound
2,4-D


Dalapon


2,4-DB


Dicamba


Dichlorprop


Dinoseb

MCPA


MCPP


2,4,5-T


2,4,5-TP


Sample
Type
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
DW
MW
MW
Spike
10.9
10.1
200
23.4
23.4
468
10.3
10.4
208
1.2
1.1
22.2
10.7
10.7
213
0.5
102
2020
2020
21400
2080
2100
20440
1.1
1.3
25.5
1.0
1.3
25.0
Mean
Recovery
75
77
65
66
96
81
93
93
77
79
86
82
97
72
100
86
81
98
73
97
94
97
95
85
83
78
88
88
72
Standard
deviation
4
4
5
8
13
9
3
3
6
7
9
6
2
3
2
4
3
4
3
2
4
3
2
6
4
5
5
4
5
aAll  results based upon seven replicate analyses.  Esterification performed using
the bubbler method. Data obtained from reference 8.

DW = ASTM Type II
MW = Municipal water
                                  8150B - 18
    Revision 2
September 1994

-------
                                   FIGURE  1.
                            DIAZOMETHANE GENERATOR
                                                                  glass tubing
    nitrogen
rubber  stopper
                                               y
                     tube 1
tube 2
                                  8150B - 19
                        Revision 2
                    September 1994

-------
                       FIGURE  2.
     GAS  CHROMATOGRAM OF  CHLORINATED HERBICIDES
Column: 1.5% SP-22SO/1.9S% SP-2401 en Supateopon (100/120
Ttmptraturc: tethcrmal it 18S°C
Dtuctor: Electron Capture
          0     12345
            RETENTION TIME (MINUTES)
                       8150B  -  20
    Revision 2
September 1994

-------
                  FIGURE 3.
GAS CHROMATOGRAM OF  CHLORINATED  HERBICIDES
  Column: 1J% SP-2250/1.95% SP-2401 on Supateoport (100/120
  Program: 140°C for 6 Min, 10°C/Minutt to 200°C
  Datactor: Electron Capture
                               I
         468
         RETENTION TIME (MINUTES)
10
12
                  8150B - 21
             Revision  2
         September  1994

-------
                  FIGURE 4.
   GAS CHROMATOGRAM OF DALAPON, COLUMN 3

         Column: 0.1% 9-1000 on 10/100 Mnh CarboMk C
         fcortm: 100°C. 10«C/Min to 180°C
         Ovttetor: Eloetron Capture
   1    I	I
   0246
MITENTION TIME (MINUTES)
                  8150B  -  22
    Revision 2
September 1994

-------
                                  METHOD 81BOB
            CHLORINATED  HERBICIDES BY GAS  CHROMATOGRAPHY
    7.2.1.1
Adjust sample
 pH with HCI.
                                         Solid
                                        Sample
7.2.1.2 Extract
 sample with
 acetone and
 diethyl ether.
7.2.1.3 Extract
  twice more.
    7.2.1.4
   Combine
   extracts.
 7.2.1.5 Check
 pH of extract,
    adjust if
   necessary.
Separate layara.
  7.2.1.5
 Re-extract
and discard
  aqueous
  phase.
   7.1.1 Follow
 Method 3580 for
 extraction, using
   diethyl ether,
acidified anhydrous
•odium sulfate and
  acidified glass
      wool.
7.2.2 Proceed
    with
 hydrolysis.
    7.1.1.2 Use
     1.0 ml of
    sample for
    hydrolysis.
7.2.3 Proceed
 with eolvent
   cleanup.
                                                   7.3.1.1 Adjust
                                                     sample pH
                                                    with H2S04.
  7.3.1.2 Extract
   with diethyl
      ether.
     7.3.1.3
Extract twice more
   and combine
     extracts.
                                7.3.2 Proceed
                                     with
                                  hydrolysis.
                                                    7.3.3 Proceed
                                                     with solvent
                                                       cleanup.
                                                                V
                                    8150B  -  23
                                                         Revision  2
                                                    September  1994

-------
                                       METHOD 81SOB
                                        (Continued)
7.4.3 Assemble
 diczomethane
   bubbler;
   generate
 diazomethane.
    7.4
  Choose
 method for
esterification
7.4.2 Prepare
diazomethane
 according to
     kit
 instructions.
                              7.5 Set
                          chromatographic
                            conditions.
                            7.6 Claibrate
                            according to
                           Method 8000.
                            7.6.2 Choose
                             appropriate
                             GC column.
                              7.7 Analyze
                              by GC (refer
                              to Method
                               8000).
                                                          7.7.7 Do
                                                        interferences
                                                        prevent peak
                                                         detection?
                                                       7.7.7 Process
                                                         series of
                                                         standards
                                                      through system
                                                         cleanup.
                                        8150B -  24
                                                         Revision  2
                                                    September  1994

-------
                                  METHOD 8151

   CHLORINATED HERBICIDES BY GC USING METHYLATION OR PENTAFLUOROBENZYLATION
                  PERIVATIZATION; CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method  8151  is  a  capillary gas  chromatographic  (GC)  method for
determining certain chlorinated acid herbicides and related compounds in aqueous,
soil and waste matrices.  Specifically, Method 8151 may be used  to determine the
following compounds:
      Compound Name                                    CAS  No."


      2,4-D                                           94-75-7
      2,4-DB                                          94-82-6
      2,4,5-TP (Silvex)                               93-72-1
      2,4,5-T                                         93-76-5
      Dalapon                                         75-99-0
      Dicamba                                       1918-00-9
      Dichloroprop                                   120-36-5
      Dinoseb                                         88-85-7
      MCPA                                            94-74-6
      MCPP                                            93-65-2
      4-Nitrophenol                                  100-02-1
      Pentachlorophenol                               87-86-5


      8     Chemical Abstract Services Registry Number.

      Because these compounds are produced and used in various forms  (i.e., acid,
salt, ester,  etc.), Method 8151  describes a hydrolysis step that can be used to
convert  herbicide  esters  into the acid form prior to analysis.  Herbicide esters
generally have a half-life of less than one week in soil.

      1.2   When Method 8151  is used to  analyze  unfamiliar samples,  compound
identifications  should  be supported  by  at  least  one  additional  qualitative
technique.   Sec.  8.4  provides gas  chromatograph/mass spectrometer  (GC/MS)
criteria   appropriate   for    the   qualitative   confirmation   of   compound
identifications.

      1.3   The estimated detection limits for  each of the compounds in aqueous
and soil matrices  are listed in Table 1.   The  detection limits  for a specific
waste sample may differ from those listed,  depending  upon the nature  of  the
interferences and the sample matrix.
                                   8151 - 1                         Revision 0
                                                                September 1994

-------
      1.4   The following compounds may also be determined using this method:
      Compound Name                               CAS No."


      Acifluorfen                               50594-66-6
      Bentazon                                  25057-89-0
      Chloramben                                  133-90-4
      DCPA diacid"                               2136-79-0
      3,5-Dichlorobenzoic acid                     51-36-5
      5-Hydroxydicamba                           7600-50-2
      Picloram                                   1918-02-1


      8      Chemical  Abstract Services Registry Number.

      b      DCPA monoacid and diacid metabolites  included in method scope; DCPA
            diacid metabolite used for validation studies.   DCPA is a dimethyl
            ester.


      1.5   This method  is  restricted to use  by or under the  supervision  of
analysts  experienced  in  the use  of  gas  chromatography  and  skilled in  the
interpretation of gas  chromatograms.   Each analyst must demonstrate the ability
to generate acceptable results with this method.

      1.6   Only experienced analysts should be  allowed to work with diazomethane
due to the potential hazards associated with its  use (explosive, carcinogenic).


2.0   SUMMARY OF METHOD

      2.1   Method  8151   provides   extraction,   derivatization,    and   gas
chromatographic conditions  for the analysis of chlorinated  acid herbicides  in
water, soil, and waste samples.   An option for  the hydrolysis of esters is also
described.

            2.1.1 Water  samples   are  extracted  with  diethyl  ether  and  then
      esterified  with  either diazomethane  or  pentafluorobenzyl  bromide.   The
      derivatives are determined  by gas chromatography with an electron capture
      detector  (GC/ECD).  The results are reported as acid equivalents.

            2.1.2 Soil  and  waste  samples  are  extracted  and  esterified  with
      either  diazomethane or pentafluorobenzyl  bromide.  The  derivatives are
      determined  by  gas chromatography  with  an  electron  capture  detector
      (GC/ECD).  The results  are  reported as acid equivalents.

            2.1.3 If herbicide esters are to  be determined  using this method,
      hydrolysis  conditions  for  the  esters  in water  and  soil  extracts are
      described.
                                   8151 - 2                         Revision 0
                                                                September 1994

-------
      2.2   The sensitivity of Method 8151 depends on the  level of interferences
in addition to  instrumental  limitations.   Table 1  lists  the  GC/ECD and GC/MS
detection limits that can be obtained in  aqueous and soil matrices in  the absence
of interferences.  Detection  limits for a typical  waste sample  should be higher.


3.0   INTERFERENCES

      3.1   Refer to Method 8000.

      3.2   Method  interferences  may be  caused  by contaminants  in solvents,
reagents, glassware, and other sample processing  hardware  that lead  to discrete
artifacts or elevated baselines in gas chromatograms.   All these materials must
be routinely demonstrated to be free  from  interferences under the conditions of
the analysis,  by analyzing reagent blanks.

            3.2.1 Glassware must be scrupulously cleaned.  Clean each piece of
      glassware  as  soon as  possible after  use  by rinsing  it  with  the  last
      solvent used in it.   This should be followed by detergent washing with hot
      water and  rinses  with  tap water,  then with  organic-free  reagent water.
      Glassware  should  be solvent-rinsed  with  acetone  and  pesticide-quality
      hexane.   After rinsing and drying, glassware should be sealed and stored
      in  a  clean  environment  to prevent  any  accumulation  of  dust  or  other
      contaminants.   Store glassware inverted  or  capped with  aluminum  foil.
      Immediately prior to  use, glassware should be rinsed with the next solvent
      to be used.

            3.2.2 The use of high purity reagents and solvents  helps  to minimize
      interference problems.  Purification of solvents by distillation in all-
      glass systems may be required.

      3.3   Matrix  interferences  may   be caused   by  contaminants  that  are
coextracted from the sample.   The extent of  matrix  interferences will  vary
considerably from waste  to  waste, depending upon the nature and diversity of the
waste being sampled.

      3.4   Organic acids,  especially chlorinated acids,  cause the most direct
interference  with  the   determination   by  methylation.    Phenols,  including
chlorophenols, may also  interfere with this procedure.  The determination using
pentafluorobenzylation is more sensitive,  and more prone  to  interferences from
the presence of organic acids or phenols than by methylation.

      3,5   Alkaline hydrolysis and  subsequent extraction  of the basic solution
removes many chlorinated hydrocarbons and  phthalate  esters that might otherwise
interfere with the electron capture analysis.  However,  hydrolysis may result in
the loss  of dinoseb and the  formation  of aldol  condensation products  if any
residual acetone remains from the extraction  of solids.

      3.6   The  herbicides,  being  strong  organic  acids,  react readily  with
alkaline substances and  may be lost during analysis.  Therefore,  glassware must
be acid-rinsed and then rinsed to constant pH with organic-free  reagent water.
Sodium sulfate must be acidified.
                                   8151 - 3                         Revision 0
                                                                September 1994

-------
      3.7   Sample extracts  should  be dry  prior  to methylation or  else poor
recoveries will be obtained.
4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph

            4.1.1 Gas  chromatograph  -  Analytical  system  complete  with  gas
      chromatograph suitable for Grob-type  injection  using capillary columns,
      and all  required accessories  including  detector,  capillary  analytical
      columns, recorder, gases, and syringes. A data system for measuring peak
      heights and/or peak areas is recommended.

            4.1.2 Columns

                  4.1.2.1     Narrow Bore Columns

                        4.1.2.1.1   Primary  Column  1  -  30  m x  0.25 mm,  5%
                  phenyl/95%   methyl   silicone  (DB-5,  J&W   Scientific,   or
                  equivalent),  0.25 yum film thickness.

                        4.1.2.1.2   Primary Column la (GC/MS)  - 30 m x 0.32 mm,
                  5%  phenyl/95%  methyl  silicone,   (DB-5,  J&W Scientific,  or
                  equivalent),  1  |um film thickness.

                        4.1.2.1.3   Column  2   -  30  m  x 0.25  mm DB-608  (J&W
                  Scientific or equivalent) with a 25 fj,m film thickness.

                        4.1.2.1.4   Confirmation Column -  30 m x  0.25 mm,  14%
                  cyanopropyl  phenyl  silicone,  (DB-1701,  J&W Scientific,  or
                  equivalent),  0.25 jttm film thickness.

                  4.1.2.2     Wide-bore Columns

                        4.1.2.2.1   Primary Column - 30 m x 0.53 mm DB-608 (J&W
                  Scientific or equivalent) with 0.83 jum film thickness.

                        4.1.2.2.2   Confirmation Column -  30 m x  0.53 mm,  14%
                  cyanopropyl  phenyl  silicone,  (DB-1701,  J&W Scientific,  or
                  equivalent),  1.0 urn film thickness.

            4.1.3 Detector - Electron Capture Detector (ECD)

      4.2   Kuderna-Danish (K-D)  apparatus

            4.2.1 Concentrator tube - 10 ml graduated (Kontes K-570050-1025 or
      equivalent).  A  ground glass stopper is used  to  prevent  evaporation of
      extracts.

            4.2.2 Evaporation  flask   -   500   mL  (Kontes  K-570001-500   or
      equivalent).   Attach  to  concentrator  tube  with  springs,  clamps,  or
      equivalent.


                                   8151 - 4                         Revision 0
                                                                September 1994

-------
            4.2.3 Snyder  column  - Three  ball  macro  (Kontes  K-503000-0121  or
      equivalent).

            4.2.4 Snyder  column   -  Two  ball  micro  (Kontes  K-569001-0219  or
      equivalent).

            4.2.5 Springs - 1/2 inch (Kontes K-662750 or equivalent).

      4.3   Diazomethane Generator:   Refer to Sec. 7.5 to determine which method
of diazomethane generation should be used for a particular generation.

            4.3.1 Diazald Kit - Recommended for the generation of diazomethane
      (Aldrich Chemical Co., Cat No. 210,025-0, or equivalent).

            4.3.2 As  an  alternative, assemble  from  two 20  mm x 150  mm test
      tubes,  two  Neoprene  rubber stoppers,  and  a source  of nitrogen.   Use
      Neoprene rubber stoppers with holes drilled in them to accommodate glass
      delivery  tubes.    The exit  tube must  be drawn  to  a  point  to  bubble
      diazomethane through the sample extract.   The generator  assembly is shown
      in Figure 1.  The procedure  for use  of this type of generator is given in
      Sec. 7.5.

      4.4   Other Glassware

            4.4.1 Beaker - 400 mL, thick walled.

            4.4.2 Funnel - 75 mm diameter.

            4.4.3 Separatory funnel  - 500 ml,  with Teflon stopcock.

            4.4.4 Centrifuge bottle - 500 ml (Pyrex 1260 or equivalent).

            4.4.5 Centrifuge bottle - 24/40 500 ml

            4.4.6 Continuous Extractor (Hershberg-Wolfe  type,  Lab Glass No. LG-
      6915, or equivalent)

            4.4.7 Pipet - Pasteur, glass, disposable (140 mm x 5 mm ID).

            4.4.8 Vials -  10 ml,  glass, with Teflon lined screw-caps.

            4.4.9 Volumetric flasks, Class A - 10 ml to 1000 mL.

      4.5   Filter paper - 15 cm diameter (Whatman No. 1 or equivalent).

      4.6   Glass Wool - Pyrex, acid washed.

      4.7   Boiling    chips    -     Solvent    extracted    with    methylene
chloride,approximately 10/40 mesh  (silicon carbide or equivalent).

      4.8   Water  bath  -  Heated,  with  concentric  ring   cover,  capable  of
temperature control (± 2°C).   The  bath  should  be used  in  a  hood.
                                   8151 - 5                         Revision 0
                                                                September 1994

-------
      4.9   Balance - Analytical, capable of accurately weighing to 0.0001 g.

      4.10  Centrifuge.

      4.11  Ultrasonic preparation - A horn-type device equipped with a titanium
tip, or a device that will give equivalent performance, shall be used.

            4.11.1      Ultrasonic Disrupter - The disrupter must have  a minimum
      power wattage of 300 watts, with pulsing capability.   A device designed to
      reduce the  cavitation sound is  recommended.    Follow the manufacturers
      instructions for preparing the disrupter for extraction of samples.  Use
      a 3/4" horn for most samples.

      4.12  Sonabox -  Recommended with above disrupters for decreasing cavitation
sound (Heat Systems - Ultrasonics, Inc., Model 432B or equivalent).

      4.13  pH paper.

      4.14  Silica gel cleanup  column  (Bond Elut™ - Analytichem, Harbor City, CA
            or equivalent).

5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be  used  in all  tests.  Unless
otherwise  indicated,  it  is  intended  that all  reagents  shall conform  to the
specifications of the Committee on Analytical  Reagents  of the American Chemical
Society, where  such  specifications  are available.   Other  grades  may  be used,
provided it is  first ascertained that the reagent is  of sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.  All  references to  water in this method
refer to organic-free water, as defined in Chapter One.

      5.3   Sodium hydroxide  solution  (0.1  N),  NaOH.  Dissolve  4 g  NaOH in
organic-free reagent water and dilute to 1.0 L.

      5.4   Potassium  hydroxide  solution (37%  aqueous solution  (w/v)),  KOH.
Dissolve 37  g  potassium  hydroxide pellets  in organic-free  reagent  water and
dilute to 100 ml.

      5.5   Phosphate buffer pH = 2.5 (0.1 M).  Dissolve 12 g sodium phosphate
(NaH2P04) in organic-free reagent water and dilute to 1.0 L.   Add phosphoric acid
to adjust the pH to 2.5.

      5.6   N-methyl-N-nitroso-p-toluenesulfonamide  (Diazald).    High purity,
available from Aldrich Chemical Co. or equivalent.

      5.7   Silicic acid, H2Si05.   100 mesh  powder,  store  at 130°C.

      5.8   Potassium carbonate, K2C03.

      5.9   2,3,4,5,6-Pentafluorobenzyl bromide  (PFBBr),  C6F5CH2Br.   Pesticide
quality or equivalent.


                                   8151 - 6                         Revision 0
                                                                September 1994

-------
      5.10  Sodium sulfate (granular, acidified, anhydrous), Na2S04.   Purify  by
heating at  400°C for 4  hours in a shallow tray, or  by precleaning the  sodium
sulfate with  methylene chloride.   If the  sodium  sulfate  is  precleaned with
methylene chloride,  a method  blank must be analyzed, demonstrating that there  is
no interference  from the sodium sulfate.   Acidify by  slurrying 100 g  sodium
sulfate with enough  diethyl  ether  to  just  cover the solid; then add  0.1 ml  of
concentrated sulfuric acid and mix thoroughly.  Remove  the  ether under vacuum.
Mix  1  g  of the  resulting solid with  5  ml of  organic-free reagent water and
measure the pH of the mixture.   It must be below a pH of 4.   Store the  remaining
solid at 130°C.
      5.11  Solvents

            5.11.1
      equivalent.

            5.11.2

            5.11.3

            5.11.4

            5.11.5      Diethyl   Ether,   C2H5OC2H5.      Pesticide  quality  or
      equivalent.  Must  be free of peroxides as  indicated  by test strips  (EM
      Quant, or equivalent).  Procedures for removal of peroxides  are provided
      with the test strips.  After cleanup,  20 ml  of ethyl alcohol  preservative
      must be added to each liter of ether.
                        Methylene  chloride,  CH2C12.    Pesticide  quality  or


                        Acetone, CH3COCH3.   Pesticide quality or equivalent.

                        Methanol, CH3OH.  Pesticide quality or equivalent.

                        Toluene, C6H5CH3.   Pesticide  quality  or  equivalent.
                        Isooctane,  (CH3)3CH2CH(CH3)2.    Pesticide  quality  or


                        Hexane, C6H14.   Pesticide  quality or  equivalent.

                        Ethanol, absolute. C2H5OH

                        Carbitol    (diethylene    glycol    monoethyl   ether),
            5.11.6
      equivalent.

            5.11.7

            5.11.8

            5.11.9
      C2H5OCH2CH2OCH2CH20 -  optional  for producing  alcohol-free diazomethane.

      5.12  Stock standard  solutions  (1000 mg/L)  - Can  be  prepared from pure
standard materials or can be purchased as certified solutions.

            5.12.1      Prepare stock standard solutions  by accurately weighing
      about 0.010 g of  pure  acid.   Dissolve the material in pesticide quality
      acetone and dilute to  volume in a 10 ml volumetric flask.  Stocks prepared
      from pure  methyl  esters  are  dissolved in  10%  acetone/isooctane (v/v).
      Larger volumes may be  used at the convenience of the analyst.   If compound
      purity is  certified  at 96% or greater,  the weight may  be  used without
      correction to calculate the concentration of the stock standard.

            5.12.2      Transfer the  stock  standard  solutions to  vials  with
      Teflon lined  screw-caps.   Store at  4°C, protected  from light.   Stock
      standard solutions should be checked frequently for signs of degradation
                                   8151 - 7
                                                                    Revision 0
                                                                September 1994

-------
      or  evaporation,  especially immediately  prior to  preparing  calibration
      standards from them.

            5.12.3      Stock standard solutions of the derivatized acids must
      be replaced after 1 year,  or  sooner,  if  comparison with check standards
      indicates a problem.  Stock standard solutions of the free acids degrade
      more  quickly  and  should   be  replaced after  two  months,  or  sooner  if
      comparison with check standards indicates a problem.

      5.13  Internal Standard Spiking Solution (if internal standard calibration
is used) - To use this approach,  the  analyst must select one or more internal
standards that are similar in analytical  behavior  to the compounds of interest.
The  analyst  must further demonstrate  that  the  measurement  of the internal
standard is not affected  by method or matrix interferences.  The compound 4,4'-
dibromooctafluorobiphenyl   (DBOB)  has  been shown  to be  an  effective internal
standard, but other  compounds, such as 1,4-dichlorobenzene, may be used if there
is a DBOB interference.

            5.13.1      Prepare  an  internal   standard  spiking  solution  by
      accurately weighing  approximately  0.0025 g of pure DBOB.   Dissolve the
      DBOB  in acetone  and  dilute  to  volume  in  a 10  ml  volumetric  flask.
      Transfer the  internal  standard spiking solution  to a  vial  with a Teflon
      lined screw-cap, and store at  room temperature.  Addition of 10 /uL of the
      internal standard spiking  solution to 10 ml of sample extract results in
      a final  internal  standard concentration of 0.25 /^g/L.  The solution should
      be replaced if  there  is a change  in  internal standard response greater
      than 20 percent of the original  response recorded.

      5.14  Calibration standards -  Calibration standards, at  a minimum of five
concentrations  for  each  parameter  of  interest,   should  be  prepared  through
dilution  of  the stock standards with diethyl  ether  or  hexane.   One  of the
concentrations should  be at a concentration near, but above, the method detection
limit.  The remaining  concentrations should  correspond to  the  expected range of
concentrations found in real samples or should define  the working range of the
GC.   Calibration  solutions  must be replaced  after six months, or  sooner  if
comparison with check standards  indicates a problem.

            5.14.1      Derivatize each calibration standard prepared from free
      acids  in  a 10  ml   K-D  concentrator tube,  according  to  the  procedures
      beginning at Sec. 7.5.

            5.14.2      Add  a known,  constant  amount  of  one  or  more internal
      standards to each derivatized calibration standard, and dilute to volume
      with the solvent indicated in the derivative option used.

      5.15  Surrogate standards  - The  analyst should monitor the performance of
the  extraction,  cleanup   (when used),  and   determinative  step,   and  the
effectiveness of the method in dealing  with  each sample matrix, by spiking each
sample,  standard,   and  blank  with  one  or  two  herbicide   surrogates  (e.g.,
herbicides that are not  expected to be  present  in  the  sample)  recommended to
encompass the range of the temperature  program  used  in this  method.  Deuterated
analogs  of  analytes should not  be  used as surrogates  in gas chromatographic
                                   8151 - 8                         Revision 0
                                                                September 1994

-------
analysis due to coelution problems.   The  surrogate standard recommended for use
is 2,4-Dichlorophenylacetic acid (DCAA).

            5.15.1      Prepare  a  surrogate  standard  spiking  solution  by
      accurately weighing approximately 0.001 g of pure DCAA.  Dissolve the DCAA
      in acetone, and dilute to volume in a 10 ml volumetric flask.   Transfer
      the surrogate  standard  spiking  solution to a  vial with  a  Teflon lined
      screw-cap,  and  store at  room temperature.   Addition of  50 /uL  of the
      surrogate standard spiking solution to 1 L of sample, prior to extraction,
      results in a final concentration in the extract of 0.5 mg/L.

      5.16  pH Adjustment Solutions

            5.16.1      Sodium hydroxide, NaOH, 6 N.

            5.16.2      Sulfuric acid, H2S04,  12  N.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this chapter,  Organic Analytes,
Sec.  4.1.  1 L samples should be collected.

      6.2   Extracts must be stored under refrigeration  (4°C).


7.0   PROCEDURE

      7.1   Preparation of High Concentration Waste  Samples

            7.1.1 Extraction

                  7.1.1.1     Follow  Method  3580,  Waste  Dilution,  with  the
            following exceptions:

                  •     use diethyl  ether as the dilution solvent,
                  •     use acidified  anhydrous  sodium  sulfate,  and acidified
                        glass wool,
                  •     spike the sample with surrogate compound(s) according to
                        Sec. 5.16.1.

                  7.1.1.2     If the sample is to be  analyzed  for both herbicide
            esters and acids, then  the sample  extract must  be  hydrolyzed.   In
            this case, transfer 1.0 mL (a  smaller volume or  a  dilution  may be
            required if herbicide concentrations are large)  to  a 250 mL  ground
            glass Erlenmeyer flask.   Proceed to Sec.  7.2.1.8.  If the analysis
            is  for   acid   herbicides   only,   proceed   to  Sec.   7.4.5   for
            derivatization by diazomethane (if PFB derivatization is selected,
            reduce the volume  of diethyl  ether to 0.1  - 0.5 mL as per Sec. 7.4.2
            and then dilute to 4 mL  with  acetone).
                                   8151  -  9                         Revision 0
                                                                September 1994

-------
7.2   Preparation of Soil, Sediment, and Other Solid Samples

      7.2.1 Extraction

            7.2.1.1     To a  400 ml,  thick-wall  beaker  add  30  g  (dry
      weight as determined in Method 3540,  Sec.  7.2.1) of the well-mixed
      solid sample.   Adjust  the pH to  2 with concentrated hydrochloric
      acid or acidify solids  in  each beaker with 85 ml of 0.1 M phosphate
      buffer  (pH  = 2.5)  and  thoroughly mix the  contents with  a glass
      stirring rod. Spike the sample with surrogate compound(s) according
      to Sec. 5.16.1.

            7.2.1.2     The  ultrasonic  extraction  of  solids  must  be
      optimized for  each  type of sample.   In  order  for  the  ultrasonic
      extractor to efficiently extract solid samples,  the sample must be
      free flowing when the solvent  is  added.  Acidified anhydrous sodium
      sulfate should be added to clay  type  soils  (normally 1:1),  or any
      other solid that is not a free flowing sandy mixture, until  a free
      flowing mixture is obtained.

            7.2.1.3     Add 100  ml  of  methylene  chloride/acetone  (1:1
      v/v) to the  beaker.  Perform  ultrasonic extraction  for  3  minutes,
      with output control  knob set at 10 (full power) and with mode switch
      on Pulse (pulsing energy rather than continuous energy) and percent-
      duty cycle knob set  at  50% (energy on 50% of time  and  off  50% of
      time).  Allow the solids to settle. Transfer the organic layer into
      a 500 ml centrifuge bottle.

            7.2.1.4     Ultrasonically extract the sample twice more using
      100 ml of methylene chloride  and the  same  ultrasonic conditions.

            7.2.1.5     Combine the  three organic extracts from the sample
      in the  centrifuge  bottle  and  centrifuge 10 minutes  to  settle the
      fine particles.  Filter the combined  extract  through filter paper
      (Whatman #1, or equivalent) containing 7-10  g  of  acidified sodium
      sulfate into a 500 ml 24/40 Erlenmeyer flask.  Add  10 g of acidified
      anhydrous  sodium sulfate.    Periodically,  vigorously  shake  the
      extract and  drying  agent  and  allow the drying  agent  to remain in
      contact with the extract for a minimum of 2 hours.  See NOTE  in Sec.
      7.3.1.6  that  emphasizes  the  need for  a  dry extract prior  to
      esterification.

            7.2.1.6     Quantitatively transfer  the contents of the flask
      to  a  500-mL Kuderna-Danish flask with  a  10-mL concentrator  tube
      attached.  Add  boiling  chips  and attach the macro  Snyder column.
      Evaporate the extract on the water bath to  a volume of approximately
      5 ml.  Remove the flasks from the water bath  and  allow them to cool.

            7.2.1.7     If  hydrolysis  or  additional  cleanup   is  not
      required and  the  sample  is dry,  proceed to Sec.  7.4.4  -  Nitrogen
      Blowdown.
                             8151  -  10                         Revision 0
                                                          September 1994

-------
      7.2.1.8     Use this step only if herbicide esters  in addition
to herbicide acids are to be determined:

            7.2.1.8.1   Add 5 ml  of 37% aqueous potassium hydroxide
      and 30 ml  of  water to the extract.   Add  additional  boiling
      chips to the  flask.   Reflux  the mixture on  a  water  bath at
      60-65°C  until  the hydrolysis  step  is completed (usually 1 to
      2 hours).  Remove the flasks  from the water bath and cool to
      room temperature. CAUTION - the presence of residual  acetone
      will result in the  formation of aldol  condensation products
      which will  cause GC interference.

            7.2.1.8.2   Transfer the  hydrolyzed aqueous  solution to
      a 500 ml separatory funnel  and extract the  solution  three
      times with  100 mL portions of  methylene chloride.  Discard the
      methylene chloride phase.   At this point the basic (aqueous)
      solution contains the herbicide salts.

            7.2.1.8.3   Adjust the pH of the solution  to  <2 with
      cold (4°C)  sulfuric  acid (1:3)  and  extract once with 40 ml of
      diethyl  ether  and twice with 20 ml portions  of ether.  Combine
      the extracts and  pour them  through  a pre-rinsed drying column
      containing 7 to 10 cm of acidified anhydrous sodium sulfate.
      Collect the dried extracts  in a  500 ml  Erlenmeyer flask (with
      a 24/40 joint) containing 10  g of acidified anhydrous sodium
      sulfate.    Periodically,  vigorously shake  the extract  and
      drying agent and allow the drying agent to remain in contact
      with the extract for a minimum of 2 hours.  See NOTE in Sec.
      7.3.1.6 that emphasizes the need for  a dry extract prior to
      esterification.   Quantitatively transfer the contents of the
      flask  to  a  500-mL   Kuderna-Danish   flask  with  a  10-mL
      concentrator tube attached when the extract is  known  to be
      dry.

            7.2.1.8.4   Proceed to Sec. 7.4,  Extract Concentration.
      If additional  cleanup is required,  proceed to Sec. 7.2.1.9.

      7.2.1.9     Use this step  if additional cleanup  of the non-
hydrolyzed herbicides is required:

            7.2.1.9.1   Partition the herbicides by extracting the
      methylene  chloride  from   7.2.1.7   (or diethyl  ether  from
      7.2.1.8.4)   with 3 x 15 ml portions of aqueous base prepared
      by carefully mixing 30 ml  of  reagent water into 15 ml of 37%
      aqueous  potassium hydroxide.   Discard the  methylene chloride
      or ether phase.  At  this point  the basic  (aqueous) solution
      contains the herbicide salts.

            7.2.1.9.2   Adjust the  pH of the solution  to  <2 with
      cold (4°C)  sulfuric  acid (1:3) and  extract once with 40 mL of
      diethyl  ether  and twice with 20 mL portions  of ether.  Combine
      the extracts and  pour them through  a pre-rinsed drying column
      containing  7 to 10 cm of acidified  anhydrous sodium sulfate.
                      8151  - 11                         Revision 0
                                                    September 1994

-------
            Collect the dried extracts in  a  500 ml Erlenmeyer flask (with
            a 24/40 joint) containing 10 g of acidified anhydrous sodium
            sulfate.   Periodically,  vigorously  shake  the extract  and
            drying agent and allow the drying agent to remain in contact
            with the extract for a minimum of 2 hours.  See NOTE in Sec.
            7.3.1.6 that emphasizes the need  for a  dry extract prior to
            esterification.  Quantitatively transfer the contents of the
            flask  to  a   500-mL   Kuderna-Danish   flask  with  a  10-mL
            concentrator tube  attached  when the  extract  is  known  to be
            dry.

                  7.2.1.9.3   Proceed   to  section   7.4  for   extract
            concentration.

            7.2.1.10    An alternative wrist-shaker extraction procedure
      can be found in Sec. 7.2 of Method 8150.

7.3   Preparation of Aqueous Samples

      7.3.1 Separatory Funnel

            7.3.1.1     Using  a  graduated  cylinder,  measure  out  a  1-L
      sample and  transfer it into  a  2-L  separatory funnel.   Spike  the
      sample with surrogate compound(s) according to Sec. 5.15.1.

            7.3.1.2     Add 250 g of NaCl  to the  sample, seal, and shake
      to dissolve the salt.

            7.3.1.3     Use this  step  only if herbicide esters  in addition
      to herbicide acids, are to be determined:

                  7.3.1.3.1   Add 17 ml of 6 N NaOH to the sample,  seal,
            and shake.  Check the pH of the sample with pH paper; if the
            sample does not have  a pH  greater  than or equal to 12, adjust
            the pH by adding more  6 N NaOH.   Let  the  sample sit at room
            temperature until the hydrolysis step is completed (usually 1
            to  2  hours),  shaking  the  separatory  funnel and  contents
            periodically.

                  7.3.1.3.2   Add  60  ml  of  methylene  chloride to  the
            sample  bottle  and rinse  both  the bottle  and the graduated
            cylinder. Transfer the  methylene  chloride  to the separatory
            funnel and extract  the sample by vigorously shaking the funnel
            for  2  minutes,  with  periodic  venting  to  release  excess
            pressure.  Allow the  organic layer to separate from the water
            phase for a minimum of 10  minutes.  If the emulsion interface
            between the  layers is  more  than one-third the volume of the
            solvent layer, the analyst must employ mechanical techniques
            to  complete  the  phase separation.   The  optimum technique
            depends upon the sample, but may include stirring, filtration
            through glass wool, centrifugation, or other physical methods.
            Discard the methylene chloride phase.
                             8151  -  12                         Revision 0
                                                          September 1994

-------
            7.3.1.3.3   Add  a second  60 ml  volume  of methylene
      chloride to  the  separatory  funnel  and repeat the extraction
      procedure  a  second  time, discarding  the  methylene chloride
      layer.  Perform a third extraction in the same manner.

      7.3.1.4     Add 17 ml of cold (4°C) 12 N sulfuric  acid to the
sample (or hydrolyzed sample),  seal, and shake to mix.   Check the pH
of the sample with pH paper:   if the sample  does not have a pH less
than or equal to 2, adjust the pH by adding more acid.

      7.3.1.5     Add 120 ml diethyl ether to the sample, seal, and
extract the sample by vigorously shaking the funnel for 2 min with
periodic  venting to release  excess pressure.   Allow the  organic
layer to separate from the water phase  for a minimum of  10 min.  If
the emulsion  interface  between layers  is more  than  one third the
volume of the solvent  layer,  the analyst  must  employ mechanical
techniques to complete the phase separation.  The optimum techniques
to complete the  phase  separation  depends  upon  the  sample,  but may
include stirring, filtration through glass wool, centrifugation, or
other  physical   methods.    Remove  the aqueous  phase  to   a  2  L
Erlenmeyer flask and collect the ether  phase in  a 500 mL Erlenmeyer
flask containing approximately 10  g of acidified anhydrous sodium
sulfate.    Periodically, vigorously shake  the extract  and drying
agent.

      7.3.1.6     Return the aqueous phase to the separatory funnel,
add 60 ml of diethyl  ether to the  sample, and repeat the extraction
procedure  a second  time,  combining the  extracts  in the  500  ml
Erlenmeyer  flask.   Perform a third extraction with  60 ml  diethyl
ether in  the  same  manner.   Allow  the extract  to remain in contact
with the sodium  sulfate for approximately 2 hours.

      NOTE:       The  drying step  is  very  critical   to  ensuring
                  complete esterification.   Any moisture remaining
                  in  the  ether  will   result  in  low  herbicide
                  recoveries.   The amount  of  sodium  sulfate  is
                  adequate  if  some  free  flowing  crystals  are
                  visible when swirling the flask.   If all of the
                  sodium sulfate  solidifies in  a cake,  add  a few
                  additional grams of  acidified sodium sulfate and
                  again test by swirling.   The  2  hour drying time
                  is a  minimum, however, the  extracts may  be held
                  in contact with  the  sodium sulfate overnight.

      7.3.1.7     Pour  the dried  extract through  a funnel  plugged
with acid  washed glass wool,  and  collect  the extract  in  the K-D
concentrator.  Use  a  glass rod to crush any  caked  sodium  sulfate
during the transfer.   Rinse the Erlenmeyer flask and funnel  with 20
to 30 ml  of diethyl  ether to complete the  quantitative transfer.
Proceed to Sec.  7.4 for extract concentration.
                       8151  -  13                         Revision 0
                                                    September 1994

-------
7.4   Extract Concentration

      7.4.1 Add one or two clean boiling chips to the flask and attach a
three ball Snyder column.  Prewet the Snyder column by adding about 1 ml
of diethyl ether to the top of the column.  Place the K-D apparatus on a
hot water bath (15-20°C above the  boiling point of the solvent)  so that the
concentrator tube is partially immersed  in  the  hot water and the entire
lower rounded surface of the flask is bathed with hot vapor.  Adjust the
vertical position of the apparatus and the water  temperature, as required,
to complete the  concentration  in 10-20 minutes.   At  the proper  rate of
distillation  the balls  of the  column  will  actively chatter,  but  the
chambers will  not flood.  When the apparent volume of liquid  reaches 1 ml,
remove the K-D apparatus  from  the water bath and  allow  it  to drain and
cool  for at least 10 minutes.

      7.4.2 Remove  the Snyder  column and rinse the  flask  and its lower
joints  into  the  concentrator tube  with  1-2 mL  of diethyl ether.   The
extract  may  be  further  concentrated by  using  either the  micro Snyder
column technique  (Sec. 7.4.3) or nitrogen blowdown technique  (Sec. 7.4.4).


      7.4.3 Micro Snyder Column Technique

            7.,4.3.1     Add another one or two clean  boiling chips to the
      concentrator tube and attach a two ball micro Snyder column.  Prewet
      the column  by adding  about  0.5 ml of diethyl ether to  the top of the
      column.  Place the  K-D apparatus in a hot  water  bath so that the
      concentrator  tube is  partially  immersed  in the hot water.   Adjust
      the vertical position of the apparatus and the water temperature, as
      required,  to  complete  the  concentration in  5-10 minutes.   At the
      proper rate of distillation the  balls  of  the column will actively
      chatter, but the chambers will  not  flood.   When  the apparent volume
      of liquid  reaches 0.5  ml,  remove  the  K-D  apparatus from the water
      bath and allow it to  drain  and cool for at  least 10  minutes.  Remove
      the  Snyder column  and rinse  the  flask and  its  lower joints with
      about 0.2  ml  of diethyl  ether  and add to  the concentrator tube.
      Proceed to Sec. 7.4.5.

      7.4.4 Nitrogen Blowdown Technique

            7.4.4.1     Place the concentrator tube  in a warm water bath
      (approximately  35°C)   and  evaporate  the  solvent  volume  to  the
      required  level   using a   gentle  stream  of  clean,   dry  nitrogen
      (filtered  through a column of activated carbon).

            CAUTION:    Do not use plasticized tubing between the carbon
                        trap and the sample.

            7.4.4.2     The internal wall of the tube must  be  rinsed down
      several  times with  diethyl  ether  during  the  operation.   During
      evaporation,  the  solvent  level  in the tube  must  be  positioned to
      prevent water from  condensing into the sample (i.e., the solvent
      level should  be below  the  level  of the water bath).   Under normal


                             8151 -  14                        Revision 0
                                                          September 1994

-------
            operating conditions, the extract  should  not  be  allowed to become
            dry.  Proceed to Sec. 7.4.5.

            7.4.5 Dilute  the  extract with  1  mL  of isooctane  and  0.5 ml  of
      methanol.    Dilute  to a final  volume  of 4  mL with  diethyl ether.   The
      sample is  now ready for methylation with diazomethane.  If PFB derivation
      is being performed, dilute to 4 ml with acetone.

      7.5   Esterification - For diazomethane derivatization proceed with Sec.
7.5.1.  For PFB  derivatization proceed with  Sec.  7.5.2.

            7.5.1 Diazomethane Derivatization - Two  methods may be used for the
      generation  of  diazomethane:    the bubbler   method  (see Figure  1),  Sec.
      7.5.1.1, and the Diazald kit method,  Sec. 7.5.1.2.
            CAUTION:    Diazomethane  is  a  carcinogen
                        certain conditions.
                         and  can explode  under
            The  bubbler  method  is  suggested  when  small  batches of  samples
      (10-15)  require  esterification.   The  bubbler  method  works well  with
      samples that have low concentrations of herbicides  (e.g., aqueous samples)
      and is  safer to use than  the Diazald  kit  procedure.   The  Diazald  kit
      method is  good  for  large quantities of  samples needing esterification.
      The Diazald kit method is more effective than the bubbler method for soils
      or  samples  that may  contain  high  concentrations of herbicides  (e.g.,
      samples  such as  soils   that  may  result  in  yellow extracts  following
      hydrolysis  may  be  difficult  to  handle by the   bubbler  method).    The
      diazomethane derivatization (U.S.EPA,  1971)  procedures,  described below,
      will react efficiently with all of the  chlorinated  herbicides described in
      this method and should be used  only by  experienced  analysts,  due to  the
      potential  hazards associated  with its  use.   The  following precautions
      should be taken:

                  Use a safety screen.
                  Use mechanical  pipetting aides.
                  Do not heat  above  90°C - EXPLOSION may result.
                  Avoid grinding surfaces, ground-glass joints, sleeve bearings,
                  and glass stirrers - EXPLOSION may result.
                  Store away from alkali metals -  EXPLOSION may result.
                  Solutions of  diazomethane decompose rapidly in the presence of
                  solid materials such as copper powder, calcium chloride,  and
                  boiling  chips.
                  7.5.1.1
            (see Figure 1).
Bubbler method - Assemble the diazomethane bubbler
                        7.5.1.1.1   Add 5 mL of diethyl ether to the first test
                  tube.   Add 1  mL of diethyl ether, 1 mL of carbitol, 1.5 mL of
                  37% KOH,  and 0.1-0.2  g of  Diazald  to the second  test  tube.
                  Immediately place the exit  tube into the concentrator tube
                  containing  the  sample  extract.     Apply   nitrogen   flow
                  (10 mL/min) to bubble diazomethane  through  the  extract  for
                  10 minutes or until the yellow color of diazomethane persists.
                                  8151 - 15
                                      Revision 0
                                  September 1994

-------
      The amount of Diazald used is sufficient for esterification of
      approximately three sample extracts.  An additional 0.1-0.2 g
      of  Diazald may  be  added  (after  the  initial  Oiazald  is
      consumed) to extend the generation of the diazomethane.  There
      is sufficient  KOH present  in  the original solution to perform
      a maximum of approximately 20 minutes of total esterification.

            7.5.1.1.2   Remove  the  concentrator  tube and  seal  it
      with a Neoprene or  Teflon  stopper.  Store at room temperature
      in a hood for 20  minutes.

            7.5.1.1.3   Destroy any unreacted di azomethane by adding
      0.1-0.2 g of silicic  acid  to  the concentrator tube.  Allow to
      stand until the evolution of nitrogen gas has stopped.  Adjust
      the  sample volume  to  10.0  ml with hexane.   Stopper  the
      concentrator tube or  transfer 1 ml of sample to  a GC vial, and
      store refrigerated  if further processing will not be performed
      immediately.  Analyze by gas chromatography.

            7.5.1.1.4   Extracts should be stored at 4°C away from
      light.  Preservation  study results indicate  that most analytes
      are stable  for 28  days; however,  it  is  recommended that the
      methylated  extracts  be  analyzed  immediately  to minimize the
      trans-esterification and  other potential reactions that may
      occur.

      7.5.1.2     Diazald  kit method -  Instructions  for preparing
diazomethane are provided with the generator kit.

            7.5.1.2.1   Add 2 ml of diazomethane solution  and let
      the sample stand  for  10 minutes with occasional  swirling.  The
      yellow  color  of  diazomethane  should be evident  and should
      persist for this  period.

            7.5.1.2.2   Rinse the inside wall of the ampule with 700
      /xL   of   diethyl   ether.     Reduce   the  sample   volume  to
      approximately 2 ml to remove excess diazomethane by allowing
      the solvent to evaporate  spontaneously  at  room temperature.
      Alternatively, 10 mg of silicic acid can be added to destroy
      the excess diazomethane.

            7.5.1.2.3   Dilute the  sample  to  10.0  ml with hexane.
      Analyze  by  gas chromatography.   It  is  recommended that the
      methylated  extracts  be  analyzed  immediately  to minimize the
      trans-esterification and  other potential reactions that may
      occur.

7.5.2 PFB Method

      7.5.2.1     Add 30 juL of  10% K2C03 and 200  fj.1  of 3% PFBBr in
acetone.  Close  the  tube with a glass  stopper and  mix on a vortex
mixer.  Heat the  tube at 60°C for 3 hours.
                       8151  -  16                        Revision 0
                                                    September 1994

-------
            7.5.2.2     Evaporate the  solution  to 0.5 ml  with a gentle
      stream of nitrogen.   Add 2 ml  of  hexane and  repeat evaporation just
      to dryness at ambient temperature.

            7.5,2.3     Redissolve the residue in 2 ml of toluene:hexane
      (1:6) for column cleanup.

            7.5.2.4     Top a  silica column (Bond  Elut™  or equivalent)
      with 0.5 cm of anhydrous sodium sulfate.   Prewet the column with 5
      ml hexane and  let  the solvent drain to the top of  the  adsorbent.
      Quantitatively transfer  the reaction residue to the  column  with
      several rinsings of the toluene:hexane solution (total 2 - 3 ml).

            7.5.2.5     Elute the column with sufficient toluene:hexane to
      collect 8  ml of  eluent.    Discard  this  fraction,  which contains
      excess reagent.

            7.5.2.6     Elute  the column  with  toluene:hexane  (9:1)  to
      collect 8  ml  of eluent  containing PFB  derivatives  in a 10  ml
      volumetric flask.  Dilute to 10 ml with hexane.  Analyze  by GC/ECD.

7.6   Gas chromatographic conditions (recommended):

      7.6.1 Narrow Bore

            7.6.1.1     Primary Column 1:

            Temperature program:    60°C to 300°C,  at  4°C/mir\
            Helium carrier flow:    30 cm/sec
            Injection volume:        2 ^L,   splitless,  45 sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C

            7.6.1.2     Primary Column la:

            Temperature program:    60°C to 300°C,  at  4°C/min
            Helium carrier flow:    30 cm/sec
            Injection volume:        2 juL,   splitless,  45 sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C

            7.6.1.3     Column 2:

            Temperature program:    60°C to 300°C,  at  4°C/min
            Helium carrier flow:    30  cm/sec
            Injection volume:        2 juL,   splitless,  45 sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C
                            8151  -  17                         Revision 0
                                                          September 1994

-------
            7.6.1.4     Confirmation Column:

            Temperature program:    60°C to 300°C,  at  4°C/min
            Helium carrier flow:    30 cm/sec
            Injection volume:       2 juL,  splitless,  45  sec delay
            Injector temperature:   250°C
            Detector temperature:   320°C

      7.6.2 Wide-bore

            7.6.2.1     Primary Column:

            Temperature program:    0.5 minute at 150°C,  150°C to 270°C at
                                    5°C/min
            Helium carrier flow:    7 mL/min
            Injection volume:       1 IJ.L

            7.6.2.2     Confirmatory Column:

            Temperature program:    0.5 minute at 150°C,  150°C to 270°C at
                                    5°C/min
            Helium carrier flow:    7 mL/min
            Injection volume:       1 /LtL

7.7   Calibration

      7.7.1 The  procedure  for  internal  or external  calibration  may be
used.  Refer to Method 8000 for a description of each of these procedures.
Use Table 1 for guidance on  selecting  the  lowest  point on the calibration
curve.

7.8   Gas chromatographic analysis

      7.8.1 Refer to Method  8000.   If the internal standard calibration
technique is used, add 10 /LtL of internal  standard to the sample prior to
injection.

      7.8.2 Follow Method 8000 for instructions on  the analysis sequence,
appropriate  dilutions,  establishing daily retention  time  windows,  and
identification criteria.  Include a mid-concentration  standard after each
group of 10 samples in the analysis sequence.

      7.8.3 An  example  of a  chromatogram  for a methylated chlorophenoxy
herbicide is shown in Figure  2.   Tables 2 and 3 present retention times
for  the  target  analytes  after  esterification,  using  the  diazomethane
derivatization   procedure   and   the   PFB   derivatization   procedure,
respectively.

      7.8.4 Record the sample volume injected and the resulting peak sizes
(in area units  or peak heights).

      7.8.5 Using either the  internal  or external  calibration procedure
(Method 8000),  determine the identity  and  quantity  of  each component peak


                             8151  - 18                        Revision 0
                                                          September 1994

-------
      in the  sample  chromatogram which corresponds to  the  compounds used for
      calibration purposes.

            7.8.6 If calibration standards have been analyzed in the same manner
      as the samples (e.g. have undergone hydrolysis and esterification), then
      the calculation  of concentration given  in  Method 8000  should be used.
      However,  if  calibration is performed  using standards made  from methyl
      ester compounds (compounds not esterified by application  of this method),
      then the  calculation of concentration  must  include  a  correction for the
      molecular weight of the methyl ester versus the acid herbicide.

            7.8.7 If peak detection and  identification are prevented  due  to
      interferences, further  cleanup is required.   Before using  any cleanup
      procedure, the  analyst  must  process  a series of standards  through the
      procedure to validate elution  patterns  and  the  absence  of interferences
      from reagents.


8.0   QUALITY CONTROL

      8.1   Refer  to Chapter One  for  specific  quality  control  procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method utilized.  If extract cleanup was  performed, follow the QC
in Method 3600 and in the specific cleanup method.

      8.2   Procedures to check the GC system operation are found in Method 8000.

            8.2.1 Select a representative  spike concentration for each compound
      (acid or ester) to be measured.  Using  stock standards, prepare a quality
      control  check  sample concentrate, in  acetone,  that is  1000  times more
      concentrated than the selected concentrations.  Use this quality control
      check sample concentrate to prepare quality control  check samples.

            8.2.2 Tables 4 and 5 present bias and precision data for water and
      clay matrices, using the diazomethane derivatization procedure.  Table 6
      presents  relative  recovery  data generated using  the  PFB derivatization
      procedure and water samples. Compare the results obtained  with the results
      given in these Tables to determine if the data quality is acceptable.

      8.3   Calculate surrogate  standard  recovery on  all  standards, samples,
blanks,   and  spikes.    Determine  if  the  recovery  is  within   limits  (limits
established by performing QC  procedures outlined in  Method 8000).

            8.3.1 If recovery  is not within limits, the  following procedures are
      required:

                  8.3.1.1     Check  to  be   sure   there  are   no   errors  in
            calculations,  surrogate  solutions  and internal  standards.   Also,
            check instrument  performance.

                  8.3.1.2     Recalculate  the data and/or reanalyze the extract
            if any of the above checks reveal a problem.
                                  8151  -  19                         Revision 0
                                                                September 1994

-------
                  8.3.1.3     Reextract and reanalyze the  sample  if none of the
            above are a problem or flag the data as "estimated concentration."

      8.4   GC/MS confirmation

            8.4.1 GC/MS techniques  should  be judiciously employed  to support
      qualitative identifications made with this method.  Refer to Method 8270
      for the appropriate GC/MS operating conditions and analysis procedures.

            8.4.2 When  available,   chemical  ionization  mass  spectra may  be
      employed to aid the qualitative identification process.

            8.4.3 Should  these MS  procedures  fail   to provide  satisfactory
      results, additional  steps may  be taken before reanalysis.  These  steps may
      include the use of alternate packed or  capillary GC columns or additional
      cleanup.


9.0   METHOD PERFORMANCE

      9.1   In single laboratory studies using  organic-free reagent water and
clay/still bottom samples, the mean  recoveries presented in  Tables 4 and 5 were
obtained for diazomethane derivatization. The standard deviations of the percent
recoveries of these measurements are also in Tables 4 and 5.

      9.2   Table 6 presents relative recoveries of the target analytes obtained
using the PFB derivatization procedure with spiked water samples.


10.0  REFERENCES

1.    Fed. Reg. 1971, 38, No. 75, Pt. II.

2.    Goerlitz, D. G.; Lamar, W.L., "Determination of Phenoxy Acid Herbicides in
      Water by Electron  Capture  and Microcoulometric Gas  Chromatography,".  U.S.
      Geol.  Survey Water Supply Paper 1967, 1817-C.

3.    Burke,  J.  A.  "Gas  Chromatography  for Pesticide  Residue  Analysis;  Some
      Practical Aspects, J. Assoc. Off Anal. Chem. 1965, 48, 1037.

4.    "Extraction and Cleanup Procedures for the Determination of Phenoxy Acid
      Herbicides  in Sediment";  U.S.  Environmental  Protection  Agency.    EPA
      Toxicant and Analysis Center:  Bay St. Louis, MS,  1972.

5.    Shore, F.L.; Amick,  E.N.; Pan,  S.  T.  "Single  Laboratory Validation of EPA
      Method  8151  for  the Analysis  of Chlorinated  Herbicides  in  Hazardous
      Waste"; U.S.  Environmental  Protection  Agency.   Environmental  Monitoring
      Systems Laboratory.  Office  of Research and  Development,  Las  Vegas, NV,
      1985;  EPA-60014-85-060.

6.    Method  515.1,  "Determination of  Chlorinated  Acids  in  Water by  Gas
      Chromatography with  an  Electron  Capture  Detector",  Revision 4.0, USEPA,
      Office  of  Research  and  Development,  Environmental  Monitoring Systems
      Laboratory, Cincinnati, Ohio.

                                  8151  - 20                        Revision 0
                                                                September 1994

-------
7.    Method 1618, "Organo-halide and Organo-phosphorus Pesticides and Phenoxy-
      acid Herbicides  by Wide  Bore  Capillary Column Gas  Chromatography with
      Selective  Detectors",  Revision  A,  July  1989, USEPA,  Office  of Water
      Regulations and Standards, Washington, DC.

8.    Gurka,  D.F, Shore, F.L.,  Pan,  S-T,  "Single  Laboratory Validation of EPA
      Method  8150 for  Determination of  Chlorinated Herbicides  in Hazardous
      Waste", JAOAC, 69, 970, 1986.
                                  8151  -  21                         Revision 0
                                                                September 1994

-------
                                  Figure  1
                           DIAZOMETHANE GENERATOR
    nitrogen
                                                                  gloss tubing
rubber  stopper
                    tube I
                                           tube 2
                                 8151  -  22
    Revision 0
September 1994

-------
                                           Figure  2
                  CHROMATOGRAM OF METHYL ESTERS OF CHLOROPHENOXYACIDS
100.0-
                                                                       J
                                                                     1O46
                                                     C
                                                    813  E
                                                        866
 RIC-
 A

271
      Vj
                   393
    200
    3:20
           L
   L  3"   LI 479  64
   U**^JLJ  i  ,  I
    •       I        I " '
                                          693
          400
          6:40
  I
 600
10:OO
                                                                G
                                                               964
                                                                  H
                                                                  I
 8OO
13:20
10OO
16:40
                                        Scan Tim*
                                        A - Oalapon. methyl ester
                                        B - Oicamba, methyl ester
                                        C : MCPP, methyl estar
                                        D - MCPA. methyl ester
                                        E - Oichlorprop. methyl ester
                                        F = 2.4. D methyl ester
                                        G - Silvex. meihyl ester
                                        H - 2.4.6  T. methyl ester
                                         I = 2.4 DB. methyl ester
                                        J  Dinoseb. methyl ether
12OO
2C:OO
                                           8151 - 23
                                                                          Revision 0
                                                                     September 1994

-------
                                    TABLE  1
              ESTIMATED METHOD DETECTION LIMITS FOR METHOD 8151,
                          DIAZOMETHANE  DERIVATIZATION
Aqueous Samples




Analyte
Acifluorfen
Bentazon
Chloramben
2,4-D
Dalapon
2,4-DB
DCPA diacid"
Dicamba
3,5-Dichlorobenzoic acid
Dichloroprop
Dinoseb
5-Hydroxydicamba
MCPP
MCPA
4-Nitrophenol
Pentachlorophenol
Picloram
2,4,5-T
2,4,5-TP
GC/ECD
Estimated
Detection
Limit8
(M9/L)
0.096
0.2
0.093
0.2
1.3
0.8
0.02
0.081
0.061
0.26
0.19
0.04
0.09d
0.056d
0.13
0.076
0.14
0.08
0.075
Soil Samples
GC/ECD
Estimated
Detection
Limit"
(M9/kg)


4.0
0.11
0.12



0.38



66
43
0.34
0.16


0.28
GC/MS
Estimated
Identification
Limit0
(ng)


1.7
1.25
0.5



0.65



0.43
0.3
0.44
1.3


4.5
a  EDL = estimated detection limit;  defined  as either the MDL (40 CFR Part 136,
   Appendix B,  Revision 1.11  ),  or  a  concentration  of  analyte in  a  sample
   yielding  a  peak  in the   final   extract  with  signal-to-noise  ratio  of
   approximately 5, whichever value is higher.

b  Detection limits determined from standard solutions  corrected back to 50 g
   samples,  extracted  and  concentrated  to  10  mL,   with  5  pi  injected.
   Chromatography   using   narrow   bore  capillary  column,   0.25  fj.m   film,
   5% phenyl/95% methyl silicone.

c  The minimum amount  of analyte to give a  Finnigan INCOS  FIT  value of 800 as
   the methyl  derivative vs. the spectrum obtained  from  50 ng of the respective
   free acid herbicide.

d  40 CFR  Part 136, Appendix  B (49  FR 43234).   Chromatography  using wide-bore
   capillary column.

e  DCPA monoacid and diacid metabolites  included  in method  scope;  DCPA diacid
   metabolite used for validation  studies.   DCPA is a  dimethyl  ester.
                                   8151  -  24
    Revision 0
September 1994

-------
                                    TABLE 2
  RETENTION TIMES (MINUTES) OF METHYL DERIVATIVES OF CHLORINATED HERBICIDES
   Megabore Columns
Analyte
  Narrow Bore Columns
Primary8    Confirmation8     Primary"
Column      Column            Column
                                                         Wide-bore Columns
Confirmation"
Column
Dalapon
3,5-Dichlorobenzoic acid
4-Nitrophenol
DCAA (surrogate)
Dicamba
Dichloroprop
2,4-D
DBOB (internal std.)
Pentachlorophenol
Chloramben
2,4,5-TP
5-Hydroxydicamba
2,4,5-T
2,4-DB
Dinoseb
Bentazon
Picloram
DCPA diacidc
Acifluorfen
MCPP
MCPA
3.4
18.6
18.6
22.0
22.1
25.0
25.5
27.5
28.3
29.7
29.7
30.0
30.5
32.2
32.4
33.3
34.4
35.8
41.5


4.7
17.7
20.5
14.9
22.6
25.6
27.0
27.6
27.0
32.8
29.5
30.7
30.9
32.2
34.1
34.6
37.5
37.8
42.8






4.39
5.15
5.85



6.97

7.92
8.74





4.24
4.74




4.39
5.46
6.05



7.37

8.20
9.02





4.55
4.94
   Primary Column:
   Confirmation Column:
            Temperature program:
            Helium carrier flow:
            Injection volume:
            Injector temperature:
            Detector temperature:
   Primary Column:
   Confirmatory Column:
            Temperature program:

            Helium carrier flow:
            Injection volume:
            5%  phenyl/95% methyl  silicone
            14% cyanopropyl  phenyl  silicone
            60°C to 300°C,  at 4°C/min
            30  cm/sec
            2 nl,  splitless,  45 sec delay
            250°C
            320°C
            DB-608
            14% cyanopropyl  phenyl  silicone
            0.5 minute  at 150°C,
            150°C to 270°C,  at 5°C/min
            7 mL/min
            1 ^L
   DCPA monoacid and diacid metabolites included in method scope; DCPA diacid
   metabolite used for validation studies.  DCPA is a dimethyl ester.
                                   8151  -  25
                                            Revision 0
                                        September 1994

-------
                                    TABLE  3
    RETENTION TIMES (MINUTES) OF PFB DERIVATIVES OF CHLORINATED HERBICIDES
Herbicide
                           Gas Chromatographic Column
Thin-film DB-5a
SP-2250*
Thick-film DB-5°
Dalapon
MCPP
Dicamba
MCPA
Dichloroprop
2,4-D
Silvex
2,4,5-T
Dinoseb
2,4-DB
10.41
18.22
18.73
18.88
19.10
19.84
21.00
22.03
22.11
23.85
12.94
22.30
23.57
23.95
24.10
26.33
27.90
31.45
28.93
35.61
13.54
22.98
23.94
24.18
24.70
26.20
29.02
31.36
31.57
35.97
a  DB-5 capillary column, 0.25 urn film thickness, 0.25 mm  ID x 30 m  long.
   Column temperature, programmed: 70°C for 1 minute,  program 10°C/min.  to
   240°C,  hold for 17 minutes.

b  SP-2550 capillary column, 0.25 urn film  thickness, 0.25  mm ID x 30 m  long.
   Column temperature, programmed: 70°C for 1 minute,  program 10°C/min.  to
   240°C,  hold for 10 minutes.

c  DB-5 capillary column, 1.0 urn film thickness, 0.32 mm ID x 30 m long.
   Column temperature, programmed: 70°C for 1 minute,  program 10°C/min.  to
   240°C,  hold for 10 minutes.
                                   8151 - 26
                                                    Revision 0
                                                September 1994

-------
                               TABLE 4
               ACCURACY AND PRECISION FOR METHOD  8151
   DIAZOMETHANE DERIVATIZATION, ORGANIC-FREE REAGENT WATER MATRIX
Analyte
Acifluorfen
Bentazon
Chloramben
2,4-D
Dalapon
2,4-DB
DCPA diacidb
Dicamba
3,5-Dichlorobenzoic acid
Dichloroprop
Dinoseb
5-Hydroxydicamba
4-Nitrophenol
Pentachlorophenol
Picloram
2,4,5-TP
2,4,5-T
Spike
Concentration
(M9/L)
0.2
1
0.4
1
10
4
0.2
0.4
0.6
2
0.4
0.2
1
0.04
0.6
0.4
0.2
Mean8 Standard
Percent Deviation of
Recovery Percent Recovery
121
120
111
131
100
87
74
135
102
107
42
103
131
130
91
117
134
15.7
16.8
14.4
27.5
20.0
13.1
9.7
32.4
16.3
20.3
14.3
16.5
23.6
31.2
15.5
16.4
30.8
Mean percent recovery calculated from 7-8 determinations of spiked
organic-free reagent water.

DCPA monoacid and diacid metabolites included in method scope; DCPA
diacid metabolite used for validation studies.  DCPA is a dimethyl ester.
                             8151  -  27
    Revision 0
September 1994

-------
                                  TABLE  5
                  ACCURACY  AND PRECISION FOR  METHOD  8151
                 DIAZOMETHANE DERIVATIZATION, CLAY MATRIX
Analyte
     Mean
Percent Recovery8
   Linear
Concentration
   Rangeb
   (ng/g)
      Percent
     Relative
Standard Deviation0
      (n=20)
Dicamba
MCPP
MCPA
Dichloroprop
2,4-D
2,4,5-TP
2,4,5-T
2,4-DB
Dinoseb
95.7
98.3
96.9
97.3
84.3
94.5
83.1
90.7
93.7
0.52
620
620
1.5
1.2
0.42
0.42
4.0
0.82
- 104
- 61,800
- 61,200
- 3,000
- 2,440
- 828
- 828
- 8,060
- 1,620
7.5
3.4
5.3
5.0
5.3
5.7
7.3
7.6
8.7
   Mean percent recovery calculated from 10 determinations of spiked clay
   and clay/still bottom samples over the linear concentration range.

   Linear concentration range was determined using standard solutions and
   corrected to 50 g solid samples.

   Percent relative standard deviation was calculated using standard
   solutions, 10 samples high in the linear concentration range, and 10
   samples low in the range.
                                 8151  -  28
                                               Revision 0
                                           September  1994

-------
   QJ       vocMOooococom
            en en en co ^*- en oo en
                   co       oo o co o r-. vo oo —i
                                                                                                                             O —I

                                                                                                                             in  i.
                                                                                                                            •r-  Ol
                                                                                                                             > -O
                                                                                                                             O)  E
                                                                                                                            C£  QJ
                                                                                                                               -M
                                                                                                                                Q.
                                                                                                                                0)
                                                                                                                               oo
                                                        CM
                            i—ii—icoinooi—icoio    i—i

                            cncficncoiDcncocn    co
   oo
   LU
   O
   CO
                   VO
                                        o


                                        CO
   oo
   LU
LU a

co ca
   oo
   LU
   l—(
   a:
   LU

   o
   o
   LU
                in
                a>
a>

o
o
a>
m
>
(O
                                                 1 •—i    ro
                   CO
                            CT>COCOCOU">cr>COCT>
   F"^       en       *^~

otvioomooco
ocnooiooooo
                                                        m
                                                        oo
                                                        ro
                                                        CTl
            i—i    o m    in e\j       oo


            cnoooooocnioo    en
               i—I       r—I        i—I





            cocvii^ooooooiOro    i^>


            oocnr^oocnooiocn    oo
                                                     fO
                                                     10
                                                                     S_
                                                                     ai
                                                    -o
                                                     O)
                                                                     a.
                                                                     fcO
                                         en

                                         a>
                                                                     o>
                                                                     c:
                                         to
                                         3
                                         d)
                                        -a
                                                                                                                            en
                                                                                                                            CVJ
                                                                                                                            CO
                            mi
    m a
                            Q_ rcj   »  i     c

                            (_> -I— <_) -r-  •>•!—  -  ^    QJ
                                                    C
                                                    0)
                                                    o
                                                    S-
                                                    O)
                                                    Q.

-------
                                           METHOD 8151
CHLORINATED HERBICIDES  BY  GC USING  METHYLATION  OR PENTAFLUOROBENZYLATION
                     DERIVATIZATION:  CAPILLARY  COLUMN TECHNIQUE

                  Extraction/Hydrolysis  of Waste  and  Soil  Samples
               NO
1

Concentrate and/or
dilute based on
whether denvatjzatton
is by diazomethane
orPFB
                                                   70 Does
                                                  sample con
                                                  tain a high
                                                   cone of
                                                    waste''
                         7.2.1.8.1 AddKOHand
                         water. Redux for 2 hrs.
                         Allow to cool.
                        7.2.1.8.2 Transfer the
                        hydrolyzed solution to a
                        sep funnel and extract 3
                        times with MeCI.
                        Discard extracts.
                        72.1.8.3 Acidify and
                        extract 3 times with
                        diethyl e»wr.  Combine
                        and dry the extracts 2 hrs.
      7.2.1.8.4 Proceed to
      Section 7.4 to concentrate
      extract
72.1.9.1 Extract 3 limes
withKOH. Discard the
MeCI.
                                          72.192 Acidify and
                                          extract 3 times with
                                          diethyl ether. Combine
                                          and dry the extracts 2 hrs.
                             72.1 1  Weigh sample
                             and add to beaker;
                             add acid and spike;
                             mix well.
                             7.2.1.2 Optimize
                            ultrasonic solid extrac-
                            tion for each matrix
                            72.1.3 Add MeCI/
                            acetone to sample 4
                            extracts min.;let
                            settle 4 decant extract
                            72.1445 Ultra-
                            sonically extract sample
                            2 more times with MeCI
                                                                     72.1.5 Combine organK
                                                                     extracts, centrifuge, and
                                                                     filter extract. Dry for
                                                                      2 hrs
72.1.6 Concentrate
extract to about 5 mL
with Snyder column.
                                                                  YES
                    If hydrolysis is not
                    required, proceed to Section
                    74.4, Nitrogen Slowdown
                     72.1.7
                  Does analysis
                 ndude herbicide
                     esters?
                                              8151  -  30
                                                    Revision  0
                                              September  1994

-------
                                          METHOD  8151
                                          (continued)

Extraction/Hydrolysis  of  Aqueous  Samples  and  Extract  Concentration
             73.1.1 Measure 1 Lot
             sample and transfer to
             a 2L sep. funnel.
             7.3.1.2 AddZSOgNaCI
             to sample and shake
             to dissolve
                   7.3.1 3
                Does analysis
            include heoKXJe esters''
           7.3.1.4 Add12Nsulfunc
           acid and shake.  Add
           until pH < 2
           7.3.1.5  Adddiethyl
           ether to sample and
           extract Save both
           phases
            731.3.1  AddGNNaOHto
            sample and shake.  Add
            until pH> 12. Let stand
            1 hr.
            73.13.2 AddMeCland
            extract by shaking for
            2min. Discard MeCI.
       73.1.6 Return aqueous phase
       to separatory funnel and repeat
       extraction 2 more times, combine
       extracts, and allow extract to
       remain in contact with sodium
       sulfate for 2 Mrs.
                    Does
                   difficult
                emulsion form
                 >1/3 solvent
                   volume?
Employ mechanical techniques
to complete phase separation
(e.g. stirring, filtration through
glass wool, centrifugatfon, or
other physical methods).
Discard MeCI
              73.1.3.3 Repeat
              extraction twice more.
              Discard MeCI.
                                        Employ mechanical techniques
                                        to complete phase separation
                                        (e.g. stirring, filtration through
                                        glass wool, centnfugation, or
                                        other physical methods).
                                        Save both phases.
                                                                                     73.1.7 Pour extract
                                                                                     through glass wool and
                                                                                     proceed to Section 7 4 1
                                                                                    7.4.1  Place K-D apparatus
                                                                                    in water bath, concentrate
                                                                                    and cool
742- 7.4.4 Complete
concentration with rmcro-
Snyder column or nitrogen
blow down.
                                           74.5 Dilute extract
                                           with 1 ml isooctane and
                                           0.5 ml methanol
                                           8151  -  31
                                                     Revision  0
                                               September  1994

-------
     METHOD 8151
     (continued)

Extract Derivatization



7.4.5 Dilute extract
to 4 mL with acetone
i
7.5.2.1 Add potassium
carbonate and PFBBr
Close tube, mix & heat
1
752.2 Evaporate with
nitrogen to 0.5 mL. Add
2 mL hexane and repeat
1
752.3 Redissorve the
residue in 2 mL toluene:
hexane (1 6)
1
7.52.4 Load sodium
sulfate / silica cleanup
column with residue.
1 1
7.5.2.5 Bute column
with enough toluene :
hexane to collect 8 mL
eluant
/7.4 5 Witt\
PFB / PFBordlazo- \. Diazo-
* vsmetwnedenva->me*ane
\zattonbe /
x^x
7.5.1.1 Asa
diazomethar
(Figur
i
7.5.1 1.1 Add 5(1
tube. Addl mLd
1 mL carbtot. 1 5
and 0.1 - 0.2 g Oil
2nd tube. Bubble
for 10 min or until
i
BmWethe Bubbler
Mbubbler «— — <
91)
»
nL to 1st test
iethyl ether,
mL ot 37% KOH
izaldtolhe
with nitrogen
yellow persists

7.5.1.1 2 Remove con-
centrator tube and seal
it. Store at room temp.
i
7.5.1.1.3 Addsilicii
concentrator tube a
until nitrogen evolul
stopped. Adjust sai
tolOmLwithhexa
Immediate analysis
(
caodto
nd let stand
km has
mple volume
•w Stopper.
is recommended
i
75^6 Discard 1st fraction
and continue elution with
enough toluene . hexane (1 : i
to collect 8 mL more eluant
Transfer to a 1 0 mL volumetrk
flask and dilute to the mark
with hexane
75.1 1 5 If necessary
store at 4 C in the dark
n for a max of 28 days
-
•
^ 7fi1 A 769 Satftr _^
conditions


7.4.5 Dilute extract
to 4 mL with diethyl
ether
/7.5.1\.
/ WHI the \ DiazaW
S Bubbler or *• Nj^L.
v DiazaW Kit /""""
\me«iodbe/
N-used?/
\ !
7.5.1.2 Read
instructions
\ i
7.5.1.2.1 Add2n
diazomethane sol
Let stand for 10 rr
and swirl
i

kit

nL
utton
tin

75.1.2.2 Rinse ampule with
diethyl ether and evaporate
to 2 mL to remove diazomethane
Alternatively, silicic acid
may be added.
i
75 1.2.3 Dilutes:
to 10mL with hex;



imple
me

       8151  -  32
    Revision 0
September 1994

-------
                            METHOD  8151
                            (continued)

              Analysis  by  Gas  Chromatography
    7.7 Internal or external
    calibration may be uMd
    (See method 8000).
                                 7.8.1 Add 10 uL internal
                                 standard to the sample
                                 prior to injection.
7.8.2 See method 8000 for
analysis sequence, appropriate
dilutions, establishing daily
retention time windows, and
identification en tana. Check
stds every 10 samples.
   7.8.4 Record volume
   infected and the resulting
   peak sizes.
7.8.5 Determine the
identity and quantify
        it pi
      78.6
    Have stds
   and samples
 been prepared and
   analyzed (he
        way?
Calculate the correction
for molecular weight of
methyl ester vs herbicide
                                        78.6 Calculate con-
                                        centration using procedure
                                        in Method 8000.
7.8.7 Perform further
cleanup if necessary
1

                             8151   -  33
                                     Revision  0
                               September  1994

-------
                                 METHOD 8240B

  VOLATILE ORGANIC COMPOUNDS BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY (6C/MS)


1.0   SCOPE AND APPLICATION

      1.1   Method 8240  is  used  to determine volatile organic  compounds  in a
variety of solid waste matrices.   This method  is applicable to nearly all types
of samples, regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily  wastes,  mousses,  tars,
fibrous  wastes,  polymeric  emulsions,   filter  cakes,  spent  carbons,  spent
catalysts, soils, and sediments.   The following compounds can be determined by
this method:
Appropriate Technique
Analyte
Acetone
Acetonitrile
Acrolein (Propenal)
Acrylonitrile
Allyl alcohol
Ally! chloride
Benzene
Benzyl chloride
Bromoacetone
Bromochloromethane (I.S.)
Bromodichloromethane
4-Bromofluorobenzene (surr.)
Bromoform
Bromomethane
2-Butanone (MEK)
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chlorobenzene
Chlorobenzene-d5 (I.S.)
Chi orodi bromomethane
Chloroethane
2-Chloroethanol
bis-(2-Chloroethyl) sulfide
2-Chloroethyl vinyl ether
Chloroform
Chloromethane
Chloroprene
3-Chloropropionitrile
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
CAS No.b
67-64-1
75-05-8
107-02-8
107-13-1
107-18-6
107-05-1
71-43-2
100-44-7
598-31-2
74-97-5
75-27-4
460-00-4
75-25-2
74-83-9
78-93-3
75-15-0
56-23-5
302-17-0
108-90-7
3114-55-4
, 124-48-1
75-00-3
107-07-3
505-60-2
110-75-8
67-66-3
74-87-3
126-99-8
542-76-7
96-12-8
106-93-4
Purge-and-Trap
PP
PP
PP
PP
PP
a
a
PP
PP
a
a
a
a
a
PP
PP
a
PP
a
a
a
a
PP
PP
a
a
a
a
ND
PP
a
Direct
Injection
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
pc
a
a
                                  8240B - 1
    Revision 2
September 1994

-------
               Appropriate Technique
Analyte
Dibromomethane
l,4-Dichloro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
l,2-Dichloroethane-d4(surr.)
1,1-Dichloroethene
trans- 1,2-Di chl oroethene
1,2-Dichloropropane
l,3-Dichloro-2-propanol
ci s- 1 , 3-Di chl oropropene
trans- 1,3-Di chl oropropene
1 , 2 , 3 , 4-Di epoxybutarie
1,4-Difluorobenzene (I.S.)
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Malononitrile
Methacrylonitrile
Methyl ene chloride
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Pentachloroethane
2-Picol ine
Propargyl alcohol
6-Propiolactone
Propionitrile
n-Propylamine
Pyridine
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2,2-Tetrachloroethane
Tetrachl oroethene
Toluene
Toluene-d8 (surr.)
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chl orof 1 uoromethane
CAS No.b
74-95-3
764-41-0
75-71-8
75-34-3
107-06-2
107-06-2
75-35-4
156-60-5
78-87-5
96-23-1
10061-01-5
10061-02-6
1464-53-5
540-36-3
123-91-1
106-89-8
64-17-5
100-41-4
75-21-8
97-63-2
591-78-6
78-97-7
74-88-4
78-83-1
109-77-3
126-98-7
75-09-2
74-88-4
80-62-6
108-10-1
76-01-7
109-06-8
107-19-7
57-57-8
107-12-0
107-10-8
110-86-1
100-42-5
630-20-6
79-34-5
127-18-4
108-88-3
2037-26-5
71-55-6
79-00-5
79-01-6
75-69-4
Purge-and-Trap
a
PP
a
a
a
a
a
a
a
PP
a
a
a
a
PP
i
i
a
PP
a
PP
NO
a
PP
PP
PP
a
a
a
PP
i
PP
PP
PP
PP
a
i
a
a
a
a
a
a
a
a
a
a
Direct
Injection
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
8240B - 2
    Revision 2
September 1994

-------
                                                  Appropriate Technique
                                                                  Direct
Analyte                              CAS No.b   Purge-and-Trap    Injection
1,2,3-Trichloropropane
Vinyl acetate
Vinyl chloride
Xylene (Total)
96-18-4
108-05-4
75-01-4
1330-20-7
a
a
a
a
a
a
a
a
a     Adequate response by this technique.
b     Chemical Abstract Services Registry Number.
pp    Poor purging efficiency resulting in high EQLs.
i     Inappropriate technique for this analyte.
pc    Poor chromatographic behavior.
surr  Surrogate
I.S.  Internal Standard
ND    Not determined

      1.2   Method  8240  can  be  used  to  quantitate most  volatile  organic
compounds that have boiling points below 200°C and that are  insoluble or slightly
soluble  in water.   Volatile water-soluble  compounds can  be included  in this
analytical technique.   However,  for the more  soluble compounds,  quantitation
limits are approximately ten times  higher  because  of poor purging efficiency.
The method is  also  limited to compounds that  elute  as sharp peaks  from a GC
column packed  with  graphitized carbon  lightly  coated with a carbowax.   Such
compounds  include  low molecular  weight halogenated  hydrocarbons,  aromatics,
ketones,  nitriles,  acetates, acrylates, ethers, and sulfides.   See Table 1 for
a list of compounds,  retention times,  and  their characteristic  ions  that have
been evaluated on a purge-and-trap GC/MS system.

      1.3   The  estimated  quantitation limit  (EQL)  of  Method  8240  for  an
individual compound  is approximately  5  M9/kg   (wet  weight)  for soil/sediment
samples,  0.5 mg/kg  (wet  weight)  for wastes, and 5 /^g/L for  ground  water (see
Table 2).  EQLs will be proportionately higher for sample extracts and samples
that require dilution  or reduced sample size  to avoid saturation of the detector.

      1.4   This method  is  restricted  to use by,  or  under the  supervision of,
analysts   experienced  in   the   use   of   purge-and-trap  systems  and  gas
chromatograph/mass  spectrometers,  and  skilled  in  the interpretation  of mass
spectra and their use as a quantitative tool.

      1.5   To increase purging  efficiencies  of  acrylonitrile  and  acrolein,
refer to Methods 5030 and 8030 for proper purge-and-trap conditions.


2.0  SUMMARY OF METHOD

      2.1   The volatile compounds are introduced into the gas chromatograph by
the purge-and-trap method or by direct injection (in limited applications).  The


                                   8240B -  3                        Revision 2
                                                                September 1994

-------
components are separated  via  the gas chromatograph and detected  using  a mass
spectrometer,  which   is  used  to provide  both  qualitative  and  quantitative
information.    The  chromatographic  conditions,  as  well  as  typical  mass
spectrometer operating parameters, are given.

      2.2    If the above  sample  introduction techniques  are  not applicable, a
portion of the sample is  dispersed in methanol  to dissolve the volatile organic
constituents.  A portion of the methanolic solution is combined with organic-free
reagent water in a specially designed purging chamber.  It is then analyzed by
purge-and-trap GC/MS following the normal water method.

      2.3    The  purge-and-trap process - An  inert  gas is  bubbled through the
solution  at  ambient  temperature, and the volatile  components  are efficiently
transferred  from  the aqueous  phase  to the  vapor phase.   The  vapor  is swept
through  a sorbent column  where  the  volatile  components  are trapped.   After
purging is completed,  the sorbent column is heated and backflushed with  inert gas
to  desorb  the  components  onto  a  gas  chromatographic  column.    The  gas
chromatographic column is  heated to elute the  components, which are  detected with
a mass spectrometer.


3.0   INTERFERENCES

      3.1    Interferences  purged or coextracted from the samples will  vary
considerably  from  source to  source,  depending upon the particular  sample or
extract  being  tested.  The  analytical  system, however,  should be  checked to
ensure freedom from interferences, under the analysis conditions,  by analyzing
method blanks.

      3.2    Samples  can  be  contaminated  by  diffusion  of  volatile  organics
(particularly methylene chloride  and  fluorocarbons) through the  septum  seal into
the sample during shipment and storage.  A trip blank, prepared from organic-free
reagent water and carried through the sampling and handling protocol, can serve
as a check on such contamination.

      3.3    Cross contamination  can  occur whenever  high-concentration and low-
concentration  samples  are  analyzed  sequentially.    Whenever   an  unusually
concentrated  sample   is  analyzed,  it  should be followed  by the  analysis  of
organic-free reagent  water to  check for cross contamination.  The purge-and-trap
system may  require extensive  bake-out and  cleaning  after a high-concentration
sample.

      3.4    The  laboratory  where volatile  analysis  is  performed should  be
completely free of solvents.

      3.5    Impurities in  the purge  gas and  from organic compounds out-gassing
from the  plumbing ahead  of the trap account  for the majority of contamination
problems.    The   analytical   system   must  be  demonstrated  to  be  free  from
contamination under  the  conditions  of the  analysis  by running  calibration and
reagent blanks.  The  use of non-TFE plastic coating,  non-TFE thread sealants, or
flow controllers with rubber components  in the purging device  should be avoided.
                                   8240B -  4                         Revision 2
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS

      4.1   Microsyringes - 10 /iL, 25 /*L,  100 /A, 250 /iL, 500 /xL, and 1,000 juL.
These syringes should be equipped with a 20 gauge (0.006 in. ID) needle having
a length sufficient to extend from the sample inlet to within 1 cm of the glass
frit in the purging device.   The needle length  will  depend upon the dimensions
of the purging device employed.

      4.2   Syringe valve - Two-way,  with  Luer  ends (three each), if applicable
to the purging device.

      4.3   Syringe - 5 ml_, gas-tight with shutoff valve.

      4.4   Balances  - Analytical, 0.0001 g, and top-loading, 0.1 g.

      4.5   Glass scintillation vials -  20 ml, with  screw caps and Teflon liners
or glass culture tubes with a screw cap and Teflon  liner.

      4.6   Volumetric flasks,  Class A -  10 mL and  100 ml,  with  ground-glass
stoppers.

      4.7   Vials  - 2 mL, for GC  autosampler.

      4.8   Spatula - Stainless steel.

      4.9   Disposable pipets - Pasteur.

      4.10  Heater or heated  oil  bath - Should be capable  of  maintaining the
purging chamber to within 1°C  over the temperature range of  ambient  to  100°C.

      4.11  Purge-and-trap device - The purge-and-trap device consists of three
separate pieces of equipment:  the sample  purger,  the trap,  and the desorber.
Several  complete devices are commercially  available.

            4.11.1      The  recommended purging chamber  is  designed to accept
      5 mL  samples with  a  water  column  at least  3  cm deep.   The  gaseous
      headspace between the water column and the trap must have a  total volume
      of less than 15 mL.  The purge gas must pass  through the  water column as
      finely divided  bubbles with  a  diameter of less than 3 mm at  the  origin.
      The purge gas must be  introduced  no more than  5 mm from  the  base of the
      water column.   The  sample purger, illustrated in  Figure  1,  meets these
      design criteria. Alternate  sample purge devices may be utilized, provided
      equivalent performance is demonstrated.

            4.11.2      The  trap must be at least 25 cm long and have an inside
      diameter of at least 0.105 in.   Starting  from  the  inlet,  the  trap should
      contain the following  amounts of adsorbents:  1/3 of 2,6-diphenylene oxide
      polymer,  1/3 of  silica  gel,   and  1/3  of  coconut  charcoal.    It  is
      recommended that 1.0 cm of methyl  silicone coated  packing be  inserted at
      the inlet to extend the  life of the  trap (see Figure 2).  If it  is not
      necessary to analyze  for  dichlorodifluoromethane  or other fluorocarbons
      of similar  volatility,  the charcoal  can be  eliminated and the  polymer
      increased to fill  2/3 of the trap.  If only compounds  boiling above 35°C


                                   8240B - 5                         Revision 2
                                                                September 1994

-------
are to  be  analyzed,  both the silica gel and  charcoal  can  be eliminated
and the polymer  increased to  fill  the  entire  trap.   Before initial use,
the trap should be conditioned overnight at 180°C by backflushing with an
inert gas flow of at  least 20 mL/min.  Vent  the trap effluent to the room,
not to  the analytical  column.   Prior  to daily use,  the trap  should be
conditioned for  10 minutes  at 180°C with backflushing.   The  trap may be
vented to the analytical column during daily conditioning.   However, the
column must be run through  the  temperature program prior to  analysis of
samples.

      4.11.3      The desorber should be capable of  rapidly  heating the
trap to 180°C  for desorption.  The  polymer  section of the trap should not
be heated higher than 180°C, and  the remaining sections should not exceed
220°C  during bake out mode.   The desorber design illustrated  in Figure 2
meets these criteria.

      4.11.4      The purge-and-trap device may be assembled as a separate
unit or  may be  coupled  to  a gas  chromatograph, as  shown in  Figures 3
and 4.

      4.11.5      Trap  Packing Materials

            4.11.5.1    2,6-Diphenylene  oxide  polymer  -   60/80  mesh,
      chromatographic grade (Tenax GC or equivalent).

            4.11.5.2    Methyl   silicone  packing   -   OV-1    (3%)   on
      Chromosorb-W, 60/80 mesh or  equivalent.

            4.11.5.3    Silica gel  -  35/60 mesh,  Davison, grade  15 or
      equivalent.

            4.11.5.4    Coconut charcoal -  Prepare  from Barnebey Cheney,
      CA-580-26,  lot #M-2649,  by  crushing through  26 mesh screen (or
      equivalent).

4.12  Gas  chromatograph/mass  spectrometer  system

      4.12.1      Gas chromatograph - An analytical  system complete with
a temperature programmable gas chromatograph and all required accessories
including syringes, analytical columns, and gases.

      4.12.2      Column - 6 ft x 0.1 in. ID glass, packed with 1% SP-1000
on Carbopack-B (60/80 mesh) or equivalent.

      4.12.3      Mass  spectrometer - Capable  of scanning from 35-260 amu
every 3 seconds  or less, using 70 volts (nominal) electron energy in the
electron impact  mode and producing  a  mass spectrum that  meets all the
criteria in Table 3 when 50  ng of 4-bromofluorobenzene  (BFB) are injected
through the gas  chromatograph inlet.

      4.12.4      GC/MS  interface  -  Any GC-to-MS  interface that  gives
acceptable calibration points at 50 ng or less per injection for each of
the  analytes  and achieves  all  acceptable  performance   criteria  (see


                            8240B  -  6                         Revision 2
                                                          September 1994

-------
      Table 3) may  be  used.  GC-to-MS  interfaces  constructed entirely of glass
      or of glass-lined materials are recommended. Glass can be deactivated by
      silanizing with dichlorodimethylsilane.

            4.12.5      Data  system  -   A  computer  system  that  allows  the
      continuous acquisition and storage on machine readable media of all mass
      spectra obtained  throughout the duration  of the  chromatographic program
      must  be interfaced  to the mass  spectrometer.  The  computer  must have
      software that allows searching any GC/MS data file for ions of a specified
      mass and plotting such ion abundances versus time or scan  number. This
      type  of plot  is  defined  as  an Extracted  Ion  Current  Profile  (EICP).
      Software must also be available  that  allows  integrating the abundances in
      any EICP between  specified time or  scan number  limits.  The  most recent
      version of the EPA/NIST Mass  Spectral Library should also be available.


5.0   REAGENTS

      5.1   Reagent  grade  chemicals  shall  be  used   in  all  tests.   Unless
otherwise indicated,  it  is  intended  that  all  reagents  shall  conform  to  the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where such  specifications  are  available.  Other  grades may be used,
provided it  is first ascertained that the reagent is of  sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent water -  All  references to  water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Stock solutions - Stock  solutions may be prepared from pure standard
materials or purchased  as  certified  solutions.  Prepare  stock standard solutions
in methanol, using assayed liquids  or gases, as appropriate.

            5.3.1  Place about 9.8 ml of methanol  in a 10 ml tared ground-glass-
      stoppered volumetric  flask.   Allow the  flask to  stand,  unstoppered,  for
      about 10 minutes  or until all  alcohol wetted surfaces have dried.  Weigh
      the flask to the  nearest 0.0001 g.

            5.3.2  Add  the assayed reference material,  as described below.

                   5.3.2.1    Liquids  - Using a 100 /LtL syringe, immediately add
            two or more drops of assayed reference material to the flask; then
            reweigh.   The liquid must fall directly into  the  alcohol  without
            contacting  the  neck of  the flask.

                   5.3.2.2    Gases  - To  prepare standards for  any  compounds
            that    boil   below   30°C   (e.g.    bromomethane,    chloroethane,
            chloromethane,  or vinyl chloride),  fill  a 5  mL  valved  gas-tight
            syringe with  the reference standard to the 5.0 mL mark.  Lower the
            needle  to  5 mm  above  the  methanol  meniscus.   Slowly introduce the
            reference standard above the surface of the liquid.  The heavy gas
            will  rapidly  dissolve  in the  methanol.   Standards  may  also  be
            prepared by using a lecture  bottle equipped with a Hamilton Lecture
                                   8240B  -  7                         Revision 2
                                                                September 1994

-------
             Bottle Septum (186600). Attach Teflon tubing to the side-arm relief
             valve and direct a gentle  stream of gas  into the methanol meniscus.

             5.3.3 Reweigh, dilute  to volume, stopper, and  then mix by inverting
      the flask  several  times.   Calculate the concentration in  milligrams per
      liter  (mg/L) from the net gain in weight.  When compound purity is assayed
      to  be  96%  or  greater,  the  weight  may be used  without  correction  to
      calculate the concentration of the stock standard.  Commercially prepared
      stock  standards may be used at any concentration if  they are certified by
      the manufacturer or by an  independent source.

             5.3.4 Transfer  the  stock  standard  solution into a  Teflon  sealed
      screw  cap  bottle.   Store,  with  minimal  headspace, at -10°C to -20°C and
      protect from light.

             5.3.5 Prepare  fresh  stock standards for gases weekly or sooner if
      comparison with check standards indicates a problem.  Reactive compounds
      such as 2-chloroethyl vinyl ether  and styrene may need to be prepared more
      frequently.  All other standards must be replaced  after six months.  Both
      gas and liquid  standards must be  monitored  closely  by comparison  to the
      initial calibration curve and by comparison to  QC check standards.  It may
      be  necessary  to replace the standards  more frequently if  either check
      exceeds a 20% drift.

             5.3.6 Optionally, calibration using a  certified gaseous mixture can
      be  accomplished  daily  utilizing  commercially  available  gaseous  analyte
      mixture  of bromomethane,  chloromethane,  chloroethane, vinyl  chloride,
      dichlorodifluoromethane  and  trichlorofluoromethane  in nitrogen.  These
      mixtures  of documented  quality  are  stable for as  long  as  six  months
      without refrigeration.  (VOA-CYL  III,  RESTEK Corporation,  Cat.  #20194 or
      equivalent).

      5.4    Secondary  dilution  standards  -  Using  stock  standard  solutions,
prepare in methanol,  secondary dilution standards containing  the compounds of
interest,  either singly or mixed  together.  Secondary dilution standards must be
stored with  minimal  headspace and  should  be  checked frequently  for  signs of
degradation  or  evaporation,   especially just  prior to  preparing  calibration
standards from them.

      5.5    Surrogate  standards  -  The surrogates  recommended are toluene-d8,
 4-bromofluorobenzene, and l,2-dichloroethane-d4.  Other compounds may be used
as surrogates,  depending upon the analysis  requirements.  A  stock surrogate
solution in methanol should be prepared as described  in Sec. 5.3, and a surrogate
standard spiking solution should be prepared from the stock at a concentration
of 250 M9/10 mL in methanol.  Each water sample undergoing GC/MS analysis must
be spiked with 10 juL of  the surrogate spiking  solution prior to analysis.

      5.6    Internal   standards  -  The  recommended  internal  standards  are
bromochloromethane,  1,4-difluorobenzene, and chlorobenzene-d5.  Other compounds
may be used  as internal  standards  as long as they have retention times similar
to the compounds being detected  by GC/MS.  Prepare internal standard stock and
secondary dilution standards  in methanol using the  procedures described in Sees.
5.3 and 5.4.   It is recommended  that the secondary dilution standard should be


                                   8240B -  8                        Revision 2
                                                                September 1994

-------
prepared  at  a concentration  of 25 mg/L  of each  internal  standard compound.
Addition of 10 juL of this  standard to 5.0 ml of sample or calibration standard
would be the equivalent of 50 jug/L.

      5.7   4-Bromofluorobenzene (BFB) standard - A standard solution containing
25 ng/jLtL of BFB  in methanol  should be prepared.

      5.8   Calibration standards  - Calibration standards at a minimum of five
concentrations should be prepared from the secondary dilution of stock standards
(see Sees. 5.3 and 5.4).  Prepare these solutions in organic-free reagent water.
One of  the  concentrations should  be  at  a concentration near,  but above,  the
method detection limit. The  remaining  concentrations  should correspond to the
expected range of concentrations found in real samples but should not exceed the
working range  of  the GC/MS system.   Each standard should contain each analyte for
detection by  this method.   It is  EPA's intent  that all  target  analytes for a
particular analysis be included  in the calibration standard(s).   However, these
target analytes may  not include  the entire List of Analytes  (Sec. 1.1) for which
the method has been demonstrated.  However,  the laboratory shall  not report a
quantitative result  for a target analyte that was not included in the  calibration
standard(s).  Calibration standards must be prepared daily.

      5.9   Matrix  spiking  standards  -  Matrix  spiking  standards  should  be
prepared from  volatile organic  compounds  which will  be  representative of the
compounds being investigated.  The suggested compounds are  1,1-dichloroethene,
trichloroethene,  chlorobenzene, toluene, and  benzene.   The standard should be
prepared  in  methanol, with each  compound  present   at  a concentration  of
250 jug/10.0 ml.

      5.10  Great care must  be  taken to maintain the integrity of all standard
solutions.  It is recommended that  all  standards in methanol be  stored  at -10°C
to -20°C in  screw cap  amber bottles with Teflon liners.

      5.11  Methanol,  CH3OH.   Pesticide quality or equivalent. Store apart from
other solvents.

      5.12  Reagent Tetraglyme  - Reagent  tetraglyme is defined as tetraglyme in
which interference  is  not observed at the method detection limit of compounds of
interest.

            5.12.1     Tetraglyme  (tetraethylene glycol  dimethyl ether, Aldrich
      #17, 240-5  or  equivalent), C8H1805. Purify by treatment at reduced pressure
      in a rotary evaporator. The tetraglyme should have a  peroxide content of
      less than  5  ppm as  indicated  by EM  Quant  Test Strips  (available  from
      Scientific Products Co., Catalog No. P1126-8 or equivalent).

            CAUTION:   Glycol  ethers  are  suspected carcinogens.  All  solvent
                       handling should be  done in a hood  while  using proper
                       protective  equipment to  minimize  exposure to liquid and
                       vapor.

            Peroxides  may be removed  by passing  the tetraglyme through a column
      of activated  alumina.  The tetraglyme is placed in a  round  bottom flask
                                   8240B  -  9                         Revision 2
                                                                September 1994

-------
      equipped with a standard taper joint, and the flask is  affixed to a rotary
      evaporator. The flask is immersed  in a water bath at 90-100°C and a vacuum
      is maintained  at < 10  mm  Hg for at least  two hours using a  two stage
      mechanical pump.  The  vacuum system is equipped with  an  all  glass trap,
      which  is  maintained  in  a dry ice/methanol bath.  Cool the  tetraglyme to
      ambient temperature and  add 100 mg/L of 2,6-di-tert-butyl-4-methyl-phenol
      to prevent peroxide formation. Store the tetraglyme  in  a tightly  sealed
      screw cap bottle  in an  area that is not contaminated by solvent vapors.

             5.12.2      In  order to demonstrate that all interfering volatiles
      have   been   removed   from  the  tetraglyme,   an   organic-free  reagent
      water/tetraglyme  blank must be analyzed.

      5.13   Polyethylene glycol,  H(OCH2CH2)nOH.  Free of  interferences  at the
detection limit of the  analytes.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1    See  the  introductory material  to  this  chapter, Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      Samples may be introduced  into the GC by either direct injection or purge-
and-trap procedures.  Whichever procedure is used, the instrument calibration and
sample introduction must be performed by the same procedure.

      Regardless of which sample introduction procedure is employed, establish
GC/MS operating conditions using the following recommendations as guidance.

      Recommended GC/MS operating conditions:

             Electron energy:              70 volts  (nominal).
             Mass range:                   35-260 amu.
             Scan time:                    To  give  5 scans/peak,  but not  to
                                          exceed 1  sec/scan.
             Initial column temperature:   45°C.
             Initial column holding time:  3 minutes.
             Column temperature  program:   8°C/minute.
             Final column temperature:     220°C.
             Final column holding time:    15 minutes.
             Injector temperature:         200-225°C.
             Source temperature:           According    to    manufacturer's
                                          specifications.
             Transfer line temperature:    250-300°C.
             Carrier gas:                  Hydrogen at 50 cm/sec or  helium at 30
                                          cm/sec.
      7.1   Direct  injection  -  In  very limited  applications  (e.g.  aqueous
   cess wastes), direct injection of the sample  into  the GC/MS  system with a 10
   syringe may be appropriate.  One such application  is for verification of the
                                  8240B - 10                        Revision 2
                                                                September 1994

-------
alcohol content  of an aqueous  sample prior to  determining  if the  sample is
ignitable (Methods  1010  or  1020).   In this case,  it  is  suggested  that direct
injection be used.  The detection limit is very high  (approximately 10,000 /zg/L);
therefore, it is  only permitted when concentrations in  excess of 10,000 /xg/L are
expected or for water soluble compounds that do not purge.  The system must be
calibrated by direct injection using  the  procedures described in Sec. 7.2,, but
bypassing the purge-and-trap device.

      7.2    Initial calibration for  purge-and-trap procedure

             7.2.1  Establish   the  GC/MS   operating   conditions,   using   the
      recommendations in Sec. 7.0 as guidance.

             7.2.2  Each GC/MS  system must be hardware tuned to meet the criteria
      in Table 3  for a  50 ng  injection or purging of 4-bromofluorobenzene (2 juL
      injection  of the BFB  standard).   Analyses  must  not  begin  until  these
      criteria are met.

             7.2.3  Assemble a purge-and-trap device that meets the specification
      in Sec.  4.11.   Condition the trap overnight at  180°C in the purge  mode
      with an inert gas flow  of at least 20 mL/min.  Prior to use,  condition the
      trap daily  for 10 min while backflushing at  180°C with the  column at 220°C.

             7.2.4  Connect the purge-and-trap device to a gas chromatograph.

             7.2.5  Prepare   the  final   solutions   containing   the   required
      concentrations of  calibration  standards,  including surrogate standards,
      directly in the purging device  (use freshly prepared stock solutions when
      preparing  the  calibration  standards  for  the initial calibration.)   Add
      5.0 ml of organic-free reagent  water  to the purging device.   The organic-
      free  reagent  water is  added to  the  purging device using  a 5 ml  glass
      syringe  fitted with  a  15 cm,  20  gauge needle.   The needle  is inserted
      through the sample  inlet shown  in  Figure  1.  The internal  diameter of the
      14 gauge needle that forms the sample inlet will permit  insertion of the
      20 gauge needle.  Next,  using a 10 /xL or 25 /nL microsyringe equipped with
      a long needle (Sec. 4.1), take a volume of the secondary dilution solution
      containing appropriate concentrations of  the calibration  standards (Sec.
      5.6).  Add the aliquot  of calibration solution  directly  to the organic-
      free reagent water in the purging device  by inserting  the needle  through
      the sample  inlet.  When discharging the contents of the microsyringe, be
      sure that the end of the syringe needle is well beneath the surface of the
      organic-free reagent water.  Similarly, add  10 juL of the internal standard
      solution (Sec. 5.4).   Close the 2 way syringe valve at the  sample inlet.

             7.2.6  Carry out the purge-and-trap  analysis procedure as described
      in Sec. 7.4.1.

             7.2.7  Tabulate the  area  response  of the  characteristic  ions  (see
      Table  1)  against  concentration  for each  compound and  each  internal
      standard.  Calculate response factors (RF)  for  each compound  relative to
      one of the  internal  standards.   The internal standard selected  for the
      calculation of the  RF  for a compound  should be the internal  standard  that
                                  8240B -  11                         Revision 2
                                                                September 1994

-------
has a retention time closest to the compound being measured (Sec. 7.5.2).
The RF is calculated as follows:

      RF - (AxCis)/(AisCJ

where:

      Ax    =     Area  of the characteristic ion for  the  compound being
                  measured.
      Ais    =     Area  of the characteristic ion for the specific internal
                  standard.
      Cis    =     Concentration of the  specific internal  standard.
      Cx    =     Concentration of the  compound being  measured.

      7.2.8 The average RF must be calculated for each compound using the
5  RF  values calculated  for  each compound from  the initial  (5-point)
calibration curve.  A system performance  check  should be made before this
calibration curve is used.   Five compounds (the System Performance Check
Compounds, or SPCCs) are checked for a minimum average relative response
factor.  These compounds are chloromethane, 1,1-dichloroethane, bromoform,
1,1,2,2-tetrachloroethane,   and  chlorobenzene.    The  minimum  acceptable
average  RF  for these  compounds  should  be 0.300 (>0.10  for  bromoform).
These  compounds  typically  have  RFs of  0.4-0.6 and  are  used  to check
compound instability and to check for degradation caused by contaminated
lines or active sites  in the system.  Examples of these  occurrences are:

            7.2.8.1     Chloromethane - This compound is the most likely
      compound to be lost if the  purge flow is  too fast.

            7.2.8.2     Bromoform -  This  compound is one of the compounds
      most likely to be purged very poorly if the purge flow is too slow.
      Cold spots and/or active sites in the transfer lines may adversely
      affect response.  Response of  the quantitation ion  (m/z  173)  is
      directly  affected by the  tuning  of BFB  at  ions m/z  174/176.
      Increasing  the  m/z 174/176 relative  to  m/z  95 ratio  may  improve
      bromoform response.

            7.2.8.3     Tetrachloroethane and 1,1-dichloroethane  - These
      compounds are degraded by contaminated transfer  lines in purge-and-
      trap systems  and/or active  sites in trapping materials.

      7.2.9 Using  the  RFs  from  the initial calibration,  calculate  and
record the percent relative standard deviation  (%RSD) for all compounds.
The percent RSD is calculated as  follows:

               SD
      %RSD =-_—  x 100
               RF
where:

      RSD   =     relative  standard deviation.
       RF   =     mean  of 5 initial  RFs for a compound.
      SD   =     standard  deviation of average RFs  for  a compound.


                            8240B - 12                        Revision 2
                                                          September 1994

-------
SD =


      where:
                     N  (RFj - RF)
                     I  	
                    1-1 N - 1
                  RFi    = RF for each of the 5 calibration levels
                  N     = Number of RF values (i.e., 5)

      The percent relative standard deviation should be less than 15% for
each compound.   However,  the  %RSD for each individual  Calibration Check
Compound (CCC) must be less than 30%.  Late-eluting compounds usually have
much better agreement.  The CCCs  are:

      1,1-Dichloroethene,
      Chloroform,
      1,2-Dichloropropane,
      Toluene,
      Ethyl benzene, and
      Vinyl chloride.

            7.2.9.1     If a %RSD greater  than 30 percent is measured for
      any CCC,  then corrective  action to  eliminate  a system leak and/or
      column  reactive  sites is required before reattempting calibration.

      7.2.10      Linearity -  If  the %RSD  of any compound is 15% or less,
then the  relative response factor is assumed  to  be  constant  over  the
calibration range, and the  average relative  response factor may be used
for quantitation  (Sec. 7.5.2.2).

            7.2.10.1    If the %RSD of any compound is greater than 15%,
      construct   calibration   curves  of  area  ratio   (A/Ais)   versus
      concentration using first or higher order regression fit  of the five
      calibration points.  The analyst should select the regression order
      which introduces the least calibration error into the quantitation
      (Sec. 7.5.2.4).   The use  of calibration  curves is  a recommended
      alternative  to  average  response factor calibration,  and  a useful
      diagnostic of standard preparation accuracy and absorption activity
      in the  chromatographic  system.

      7.2.11      These curves  are  verified  each  shift  by  purging  a
performance standard.   Recalibration is required only if calibration and
on-going performance criteria cannot be met.

7.3   Daily GC/MS calibration

      7.3.1 Prior to the analysis of samples, inject or purge 50 ng of the
4-bromofluorobenzene  standard.   The resultant mass  spectra  for  the  BFB
must meet all  of the criteria given  in  Table 3 before  sample  analysis
begins.   These criteria must be demonstrated each 12 hour shift.
                            8240B - 13                        Revision 2
                                                          September 1994

-------
      7.3.2  The initial calibration curve (Sec.  7.2)  for each compound of
interest must  be  checked and verified  once  every 12 hours  of analysis
time.  This  is accomplished  by  analyzing  a  calibration  standard that is
at a concentration near the midpoint concentration for the working range
of the GC/MS and checking the SPCC (Sec. 7.3.3) and CCC (Sec. 7.3.4).

      7.3.3  System   Performance  Check  Compounds  (SPCCs)  -  A  system
performance  check must be made  each  12  hours.   If the SPCC criteria are
met, a comparison of relative response factors is made for all compounds.
This is  the  same  check that is applied during  the initial  calibration.
If the minimum relative response factors are not met, the system must be
evaluated, and  corrective action must  be taken before  sample analysis
begins.  The  minimum relative response factor for volatile SPCCs is 0.300
(>0.10  for  Bromoform).   Some  possible problems  are standard  mixture
degradation,   injection port  inlet  contamination,  contamination  at the
front end  of the analytical column, and  active sites in  the  column or
chromatographic system.

      7.3.4  Calibration  Check  Compounds  (CCCs):    After  the  system
performance  check is met, CCCs listed in Sec. 7.2.9 are used to check the
validity of  the initial calibration.
      Calculate the percent drift using the following equation:

                  C - C
      % Drift = —'	—  x 100


where:
C,
      C, =    Calibration  Check  Compound  standard concentration.
      Cc =   Measured  concentration  using  selected quantitation method.

      If the percent difference for  each CCC is less  than 20%, the initial
calibration  is assumed to be valid.   If the  criterion is  not met (> 20%
drift),   for any  one  CCC,  corrective  action must  be taken.   Problems
similar to  those  listed  under  SPCCs could affect  this criterion.   If no
source of the problem can be determined after corrective action has been
taken,  a new five point  calibration MUST be  generated.   This criterion
MUST be met  before quantitative sample analysis begins.  If the CCCs are
not required analytes by  the  permit, then  all  required analytes must meet
the 20% drift criterion.

      7.3.5 The  internal standard  responses  and retention  times  in the
check calibration standard must be  evaluated  immediately after or during
data acquisition.  If the retention time for any internal standard changes
by more than 30  seconds  from the  last calibration  check (12 hours), the
chromatographic system must be  inspected for malfunctions and corrections
must  be  made,  as required.   If the  EICP area  for any of  the internal
standards changes by a factor of two (- 50% to + 100%)  from the last daily
calibration  check standard, the mass spectrometer  must be inspected for
malfunctions  and  corrections   must  be  made,  as  appropriate.    When
                            8240B - 14                        Revision 2
                                                          September 1994

-------
corrections are made,  reanalysis of samples analyzed while the system was
malfunctioning is necessary.

7.4   GC/MS analysis

      7.4.1 Water samples

            7.4.1.1     Screening of the sample  prior  to purge-and-trap
      analysis  will   provide  guidance  on whether sample  dilution  is
      necessary  and  will  prevent  contamination  of the  purge-and-trap
      system.   Two  screening  techniques  that can  be used  are:    the
      headspace  sampler (Method  3810)  using  a  gas chromatograph  (GC)
      equipped with a photo ionization  detector  (PID)  in series with an
      electrolytic conductivity detector  (HECD); and  extraction of the
      sample with hexadecane  and  analysis of the  extract  on  a GC with a
      FID and/or an ECD  (Method 3820).

            7.4.1.2     All  samples  and standard solutions must be allowed
      to warm to ambient temperature before analysis.

            7.4.1.3     Set  up the GC/MS system as outlined in Sec. 7.2.1.

            7.4.1.4     BFB  tuning criteria and  daily  GC/MS  calibration
      criteria must be met (Sec. 7.3) before analyzing samples.

            7.4.1.5     Adjust the  purge gas  (helium)  flow rate to 25-
      40 mL/min on the purge-and-trap device.   Optimize the flow rate to
      provide the best response  for  chloromethane  and bromoform, if these
      compounds are analytes.   Excessive flow rate reduces chloromethane
      response, whereas insufficient flow reduces  bromoform response (see
      Sec. 7.2.8).

            7.4.1.6     Remove the plunger from a  5 ml syringe and attach
      a closed syringe valve.   Open the sample or standard bottle,  which
      has been allowed to come to  ambient temperature, and carefully pour
      the sample  into  the  syringe barrel  to just  short  of  overflowing.
      Replace the  syringe  plunger  and  compress  the sample.    Open the
      syringe valve and vent any residual  air while adjusting the sample
      volume to 5.0 ml.  This process of taking  an  aliquot  destroys the
      validity of the  liquid  sample for future  analysis; therefore,  if
      there is only one VOA vial, the analyst should fill a second syringe
      at this time to protect  against possible loss of  sample integrity.
      This second  sample is  maintained  only until  such  time  when the
      analyst has  determined  that  the  first  sample  has been  analyzed
      properly.  Filling one 20 ml syringe would allow the use of only one
      syringe.  If a  second  analysis is needed from a syringe, it must be
      analyzed within 24 hours.   Care must be  taken to prevent  air from
      leaking into the syringe.

            7.4.1.7     The   following  procedure  is   appropriate   for
      diluting purgeable samples.   All  steps must be  performed without
      delays until the diluted sample is in a gas tight syringe.
                            8240B  -  15                         Revision  2
                                                          September  1994

-------
            7.4.1.7.1    Dilutions  may  be made in volumetric flasks
      (10 to 100 ml).  Select the volumetric flask that will  allow
      for the  necessary dilution.   Intermediate dilutions may  be
      necessary for extremely large dilutions.

            7.4.1.7.2    Calculate the approximate volume of organic-
      free  reagent water  to  be  added  to  the  volumetric  flask
      selected and add slightly less than this quantity of organic-
      free reagent water to the flask.

            7.4.1.7.3    Inject  the proper aliquot of  samples  from
      the  syringe  prepared  in  Sec.  7.4.1.6  into   the  flask.
      Aliquots of  less  than  1 ml  are  not  recommended.   Dilute the
      sample to the mark with organic-free reagent water.  Cap the
      flask, invert, and shake  three times.  Repeat above procedure
      for additional dilutions.

            7.4.1.7.4    Fill  a  5 ml syringe with the diluted sample
      as in Sec. 7.4.1.6.

      7.4.1.8     Add  10.0  /zL of  surrogate  spiking  solution  (Sec.
5.5) and 10  /iL of  internal  standard  spiking solution  (Sec.  5.6)
through the valve bore  of the  syringe;  then  close the  valve.   The
surrogate and internal  standards may be  mixed and added as a single
spiking solution.  The  addition of 10 juL  of the surrogate spiking
solution to  5  ml  of  sample  is equivalent  to  a concentration  of
50 jitg/L of each surrogate standard.

      7.4.1.9     Attach the  syringe-syringe valve  assembly to the
syringe valve on the purging device.  Open  the  syringe  valves and
inject the sample into the purging chamber.

      7.4.1.10    Close  both  valves  and  purge  the   sample  for
11.0 + 0.1  minutes at ambient temperature.

      7.4.1.11    At the conclusion of  the  purge time,  attach the
trap to the  chromatograph, adjust the device to the desorb mode, and
begin the gas  chromatographic  temperature program  and  GC/MS  data
acquisition. Concurrently,  introduce  the  trapped materials to the
gas  chromatographic  column  by  rapidly  heating  the trap to  180°C
while backflushing the trap with inert gas between 20 and 60 mL/min
for 4 minutes.   If this rapid heating requirement cannot be met, the
gas  chromatographic  column  must  be  used as  a  secondary  trap  by
cooling it to 30°C (or subambient,  if  problems  persist)  instead of
the recommended initial program temperature of 45°C.

      7.4.1.12    While the  trap  is   being  desorbed into  the gas
chromatograph,  empty the purging chamber.  Wash the chamber with a
minimum  of  two  5  ml  flushes  of organic-free  reagent  water (or
methanol followed by organic-free  reagent water) to avoid carryover
of pollutant compounds  into subsequent analyses.
                      8240B - 16                        Revision 2
                                                    September 1994

-------
             7.4.1.13    After  desorbing  the   sample   for  4  minutes,
      recondition the trap by returning the purge-and-trap device to the
      purge  mode.   Wait  15  seconds;  then  close  the syringe valve on the
      purging  device to  begin gas  flow through  the  trap.    The  trap
      temperature should be maintained at 180°C.  Trap temperatures up to
      220°C may be employed;  however,  the higher temperature will shorten
      the  useful  life of  the trap.   After approximately 7 minutes,  turn
      off  the trap heater and open the syringe valve to  stop the gas flow
      through the trap.   When cool, the trap is  ready for the next sample.

             7.4.1.14    If the initial analysis  of  a sample or a dilution
      of  the sample has  a  concentration of  analytes  that  exceeds  the
      initial calibration range, the  sample must be reanalyzed at a higher
      dilution.  Secondary ion quantitation is allowed  only when  there are
      sample  interferences  with  the  primary  ion.   When  a sample  is
      analyzed that has  saturated  ions from a compound, this analysis must
      be followed by a blank organic-free  reagent water  analysis.  If the
      blank  analysis  is not  free  of interferences,  the  system must  be
      decontaminated.   Sample analysis  may  not  resume until  a blank can
      be analyzed that  is free of  interferences.

             7.4.1.15    For matrix spike analysis, add  10 /j,L of the matrix
      spike  solution  (Sec.  5.9)  to   the  5  ml  of  sample  to  be purged.
      Disregarding  any  dilutions,  this  is  equivalent  to a concentration
      of 50  /xg/L  of each  matrix spike standard.

             7.4.1.16    All   dilutions  should  keep the  response  of  the
      major  constituents  (previously  saturated  peaks)  in  the upper  half
      of the linear range of the curve.  Proceed to Sees. 7.5.1 and 7.5.2
      for  qualitative and quantitative analysis.

      7.4.2  Water miscible liquids

             7.4.2.1     Water miscible  liquids  are   analyzed  as  water
      samples after first diluting them at least 50  fold  with organic-free
      reagent water.

             7.4.2.2     Initial  and serial dilutions  can  be  prepared  by
      pipetting  2 ml of  the sample  to  a  100  ml  volumetric flask  and
      diluting  to  volume  with organic-free  reagent water.    Transfer
      immediately to a 5 ml  gas tight syringe.

             7.4.2.3     Alternatively, prepare  dilutions directly in a 5
      ml syringe filled  with  organic-free reagent water by  adding at least
      20 /iL,  but  not  more than 100 juL of liquid sample.   The sample  is
      ready  for addition of  internal   and surrogate standards.

      7.4.3  Sediment/soil  and waste  samples -  It  is  highly  recommended
that all  samples  of this type be screened prior  to  the  purge-and-trap
GC/MS analysis.   The headspace method  (Method 3810)  or  the hexadecane
extraction  and  screening  method   (Method  3820) may  be  used  for  this
purpose.   These  samples  may  contain percent  quantities of purgeable
organics that  will  contaminate the  purge-and-trap system,  and require


                            8240B  - 17                        Revision 2
                                                          September  1994

-------
extensive cleanup  and  instrument downtime.   Use the screening  data  to
determine whether to use the low-concentration method (0.005-1 nig/kg)  or
the high-concentration method (> 1 mg/kg).

            7.4.3.1     Low-concentration method - This  is  designed for
      samples containing individual  purgeable compounds of < 1 mg/kg.  It
      is limited to sediment/soil samples and waste that  is  of a similar
      consistency (granular and  porous).  The low-concentration method is
      based on purging a heated sediment/soil sample mixed with organic-
      free reagent water containing  the surrogate and internal  standards.
      Analyze all reagent blanks and standards under the  same  conditions
      as the samples.   See  Figure 5 for an  illustration of a low soils
      impinger.

                  7.4.3.1.1    Use   a   5   g   sample   if  the  expected
            concentration  is  <  0.1  mg/kg  or a  1 g  sample  for expected
            concentrations  between  0.1 and 1 mg/kg.

                  7.4.3.1.2    The GC/MS  system  should be set up  as  in
            Sees.  7.4.1.2-7.4.1.4.   This  should be  done  prior  to the
            preparation  of  the  sample to  avoid loss of volatiles from
            standards and samples.  A heated  purge calibration curve must
            be  prepared and  used  for the quantitation  of all  samples
            analyzed  with  the  low-concentration  method.     Follow the
            initial  and  daily calibration instructions,  except  for the
            addition of  a 40°C purge temperature.

                  7.4.3.1.3    Remove the  plunger from a 5 ml Luerlock type
            syringe  equipped  with  a  syringe  valve  and   fill  until
            overflowing  with  organic-free reagent  water.   Replace the
            plunger  and  compress  the water to vent  trapped  air.   Adjust
            the  volume  to  5.0 ml.   Add 10 juL each  of surrogate spiking
            solution (Sec. 5.5)  and  internal  standard solution  (Sec. 5.6)
            to  the  syringe  through   the  valve.    (Surrogate  spiking
            solution  and   internal  standard   solution  may   be  mixed
            together.)   The addition  of 10 /uL  of  the surrogate spiking
            solution to 5 g of sediment/soil  is equivalent to 50 ^9/kg of
            each  surrogate  standard.

                  7.4.3.1.4    The sample  (for volatile organics) consists
            of  the entire  contents of  the  sample  container.    Do not
            discard  any supernatant liquids.   Mix the  contents  of the
            sample  container  with  a  narrow  metal  spatula.    Weigh the
            amount  determined  in  Sec.  7.4.3.1.1  into  a  tared  purge
            device.  Note and record the  actual weight to the nearest 0.1
            9-

                  7.4.3.1.5    Determine  the  percent dry  weight  of the
            soil/sediment sample.   This  includes waste  samples that are
            amenable to  percent  dry weight determination.   Other wastes
            should be reported on a wet-weight  basis.
                            8240B - 18                        Revision 2
                                                          September 1994

-------
                  7.4.3.1.5.1  Immediately after weighing the sample
            for extraction, weigh 5-10 g of the sample into a tared
            crucible.   Determine  the  %  dry weight of the sample by
            drying   overnight  at   105°C.    Allow  to  cool   in  a
            desiccator   before  re-weighing.     Concentrations  of
            individual  analytes  are  reported  relative to the dry
            weight  of sample.

                  WARNING:     The drying  oven  should  be contained
                              in  a  hood  or  vented.   Significant
                              laboratory  contamination  may  result
                              from a  heavily contaminated hazardous
                              waste sample.

                  % dry weight =  g  of dry sample  x 100
                                  g of sample

            7.4.3.1.6    Add the  spiked water to  the purge device,
      which contains the weighed amount of sample, and connect the
      device to the purge-and-trap system.

            NOTE:  Prior to  the attachment of  the  purge device, the
                  procedures  in Sees.  7.4.3.1.4 and 7.4.3.1.6 must
                  be performed rapidly  and without interruption to
                  avoid loss of  volatile  organics.   These  steps
                  must  be performed in a laboratory free of solvent
                  fumes.

            7.4.3.1.7    Heat  the  sample  to 40°C +  1°C and purge the
      sample for 11.0 + 0.1 minute.

            7.4.3.1.8    Proceed  with the  analysis as outlined  in
      Sees. 7.4.1.11-7.4.1.16.   Use  5 ml  of  the  same  organic-free
      reagent  water as  in  the  reagent blank.  If saturated  peaks
      occurred  or  would occur  if a 1 g  sample were analyzed, the
      high-concentration method must be followed.

            7.4.3.1.9    For low-concentration  sediment/soils  add
      10 /nL of the matrix spike solution  (Sec.  5.9) to the 5 ml of
      organic-free   reagent  water    (Sec.   7.4.3.1.3).      The
      concentration for  a  5 g  sample would be  equivalent  to  50
      jug/kg of each matrix spike standard.

      7.4.3.2     High-concentration method -  The  method  is based on
extracting  the sediment/soil  with methanol.   A waste sample  is
either  extracted   or   diluted,  depending on  its  solubility  in
methanol.   Wastes (i.e.  petroleum  and  coke   wastes)  that  are
insoluble  in   methanol  are  diluted  with reagent  tetraglyme  or
possibly polyethylene glycol  (PEG).  An  aliquot  of the  extract  is
added to organic-free reagent water containing  internal  standards.
This is  purged  at ambient temperature. All  samples with an expected
concentration of > 1.0 mg/kg should be analyzed by this method.
                      8240B -  19                         Revision  2
                                                    September 1994

-------
     7.4.3.2.1    The sample (for volatile organics)  consists
of the  entire contents  of the  sample  container.   Do  not
discard any  supernatant  liquids.  Mix  the  contents of  the
sample  container  with   a  narrow  metal  spatula.     For
sediment/soil  and  solid  wastes  that   are   insoluble   in
methanol,  weigh 4 g  (wet weight)  of sample into a tared  20 ml
vial.  Use a  top  loading balance.  Note and record the actual
weight to 0.1  gram  and  determine the percent dry weight  of
the sample using the procedure in Sec. 7.4.3.1.5.  For  waste
that is soluble  in  methanol,  tetraglyme,  or PEG, weigh  1 g
(wet weight)  into a  tared  scintillation  vial  or culture tube
or a 10 ml volumetric flask.   (If a vial  or  tube is  used,  it
must be calibrated  prior  to  use.  Pi pet  10.0 ml  of solvent
into the vial and mark the bottom of the  meniscus.  Discard
this solvent.)

     7.4.3.2.2    Quickly  add  9.0 ml  of  appropriate  solvent;
then add  1.0 ml  of  the surrogate spiking  solution to  the
vial.  Cap and shake for 2 minutes.

     NOTE: Sees.  7.4.3.2.1 and 7.4.3.2.2 must  be performed
           rapidly  and without interruption to avoid loss of
           volatile organics.  These steps must be performed
           in  a  laboratory free  from solvent  fumes.

     7.4.3.2.3    Pipet  approximately  1  ml of the  extract to
a  GC vial  for  storage,   using  a disposable  pipet.     The
remainder may be  disposed  of.  Transfer approximately 1  ml of
appropriate  solvent  to  a  separate  GC  vial   for use as  the
method blank for each set  of samples.  These extracts may be
stored at 4°C in the dark, prior to analysis.   The  addition
of  a 100  /iL  aliquot  of  each  of  these extracts  in  Sec.
7.4.3.2.6 will  give  a concentration equivalent to 6,200  M9/kg
of each surrogate standard.

     7.4.3.2.4    The  GC/MS system  should be  set  up   as  in
Sees. 7.4.1.2-7.4.1.4.   This  should  be  done  prior to  the
addition  of  the  solvent  extract to  organic-free  reagent
water.

     7.4.3.2.5    Table  4 can be used to  determine the volume
of solvent extract to  add  to the  5 ml of  organic-free reagent
water for  analysis.   If a screening procedure  was  followed
(Method 3810  or  3820),  use  the  estimated concentration  to
determine the  appropriate  volume.   Otherwise,  estimate  the
concentration range  of the sample from the low-concentration
analysis to determine the  appropriate volume.   If the sample
was  submitted  as a high-concentration  sample, start with
100 /iL.   All  dilutions  must keep the response  of the  major
constituents (previously saturated peaks) in  the  upper half
of the linear range  of the curve.
                8240B -  20                         Revision  2
                                              September 1994

-------
                  7.4.3.2.6    Remove the plunger from a  5.0  ml Luerlock
            type  syringe  equipped with a  syringe  valve and  fill  until
            overflowing with  organic-free reagent  water.    Replace  the
            plunger  and compress  the water to  vent  trapped  air.   Adjust
            the  volume  to 4.9 ml.  Pull  the  plunger back to  5.0  ml  to
            allow  volume  for  the  addition of the sample extract  and  of
            standards.  Add  10 pi of internal standard  solution.   Also
            add  the  volume  of  solvent  extract  determined  in  Sec.
            7.4.3.2.5 and a volume  of extraction  or dissolution  solvent
            to total 100 juL  (excluding  methanol in  standards).

                  7.4.3.2.7    Attach the syringe-syringe valve assembly to
            the  syringe valve on  the  purging device.   Open  the  syringe
            valve  and  inject the  organic-free  reagent  water/methanol
            sample into the purging chamber.

                  7.4.3.2.8    Proceed  with the  analysis as  outlined  in
            Sec.  7.4.1.11-7,4.1.16.   Analyze all  reagent blanks  on  the
            same  instrument as  that use for  the samples.   The standards
            and blanks should also contain 100 juL of solvent to simulate
            the sample conditions.

                  7.4.3.2.9    For  a matrix spike in the high-concentration
            sediment/soil  samples,  add 8.0  mL of  methanol,  1.0  ml  of
            surrogate spike  solution  (Sec. 5.5),  and  1.0 mL  of  matrix
            spike solution (Sec. 5.9)  as in Sec. 7.4.3.2.2.  This  results
            in a 6,200 ^g/kg  concentration of each matrix spike standard
            when  added  to  a  4 g  sample.   Add  a  100 /^L  aliquot  of this
            extract to 5 ml of organic-free reagent water for purging (as
            per Sec. 7.4.3.2.6).

7.5   Data interpretation

      7.5.1 Qualitative analysis

            7.5.1.1     The   qualitative  identification  of  compounds
      determined  by  this  method  is  based on  retention  time,  and  on
      comparison of the sample mass spectrum, after background correction,
      with  characteristic ions  in  a  reference  mass  spectrum.    The
      reference mass spectrum must be generated by  the  laboratory using
      the conditions of this  method.   The characteristic ions from  the
      reference mass  spectrum  are  defined to be the three ions of greatest
      relative intensity,  or  any ions over 30% relative intensity if less
      than three  such ions occur  in the reference  spectrum.   Compounds
      should be identified as present  when the criteria below are met.

                  7.5.1.1.1    The  intensities  of the characteristic ions
            of a compound  maximize in  the  same  scan or within one  scan  of
            each  other.   Selection of  a  peak by  a data system  target
            compound  search  routine where  the search  is  based  on  the
            presence  of  a target chromatographic  peak containing  ions
            specific  for  the target   compound at  a  compound-specific
            retention time will be accepted as meeting this criterion.
                            8240B -  21                         Revision  2
                                                          September 1994

-------
            7.5.1.1.2    The  RRT of the sample component  is  within
      + 0.06 RRT units of the RRT of the standard component.

            7.5.1.1.3    The    relative    intensities    of    the
      characteristic  ions  agree  within  30%   of   the   relative
      intensities  of  these  ions  in  the   reference   spectrum.
      (Example:   For  an ion  with an  abundance  of 50% in  the
      reference spectrum, the corresponding abundance in  a  sample
      spectrum can range between 20% and 80%.)

            7.5.1.1.4    Structural isomers that produce very similar
      mass  spectra should be  identified as individual isomers  if
      they  have   sufficiently   different  GC   retention  times.
      Sufficient GC  resolution  is achieved if  the  height of  the
      valley between two isomer peaks is  less than  25% of the  sum
      of the two peak heights.  Otherwise, structural isomers  are
      identified as isomeric pairs.

            7.5.1.1.5    Identification   is hampered  when  sample
      components are not resolved chromatographically and produce
      mass  spectra  containing  ions contributed  by  more  than  one
      analyte.  When gas chromatographic peaks obviously  represent
      more  than one sample component (i.e., a broadened  peak with
      shoulder(s)  or  a  valley  between   two  or  more   maxima),
      appropriate  selection of  analyte  spectra  and background
      spectra is important.   Examination  of extracted ion current
      profiles  of  appropriate   ions can  aid  in  the selection  of
      spectra,  and  in  qualitative identification  of compounds.
      When analytes coelute  (i.e., only one chromatographic peak is
      apparent), the identification criteria  can  be met,  but each
      analyte spectrum will  contain extraneous ions  contributed by
      the coeluting compound.

      7.5.1.2    For samples containing components  not associated
with the calibration standards, a library search may  be made for  the
purpose of tentative  identification.  The necessity to perform this
type of identification will  be  determined by  the  type of analyses
being conducted.   Guidelines  for making  tentative  identification
are:

      (1)   Relative  intensities  of major ions  in  the  reference
spectrum (ions > 10%  of  the most abundant ion)  should be  present in
the sample spectrum.

      (2)   The  relative  intensities of the major ions should agree
within ± 20%.   (Example:  For an ion with an abundance of 50% in  the
standard spectrum,  the corresponding sample ion  abundance must be
between 30 and 70%).

      (3)   Molecular ions present in the reference spectrum should
be present in the sample spectrum.
                      8240B -  22                        Revision 2
                                                    September 1994

-------
       (4)   Ions present  in  the  sample  spectrum  but  not  in  the
reference  spectrum  should  be  reviewed  for possible  background
contamination or presence of coeluting compounds.

       (5)   Ions present in the reference spectrum  but  not in the
sample spectrum should be reviewed for possible subtraction  from the
sample spectrum because of background contamination  or coeluting
peaks.  Data system library reduction  programs can sometimes create
these discrepancies.

       Computer  generated  library search  routines should  not  use
normalization  routines that  would  misrepresent  the  library  or
unknown  spectra when compared to each  other.   Only  after visual
comparison of sample with the nearest library searches will  the mass
spectral     interpretation    specialist    assign   a    tentative
identification.

7.5.2  Quantitative  analysis

       7.5.2.1     When  a  compound   has   been   identified,   the
quantitation  of that  compound  will  be  based  on the  integrated
abundance  from  the  EICP  of  the   primary  characteristic  ion.
Quantitation will take place using the internal standard technique.
The internal  standard  used shall  be  the one nearest the retention
time of that  of a given analyte (e.g. see Table  5).

       7.5.2.2     When  linearity  exists,   as  per  Sec.   7.2.10,
calculate the concentration of each identified analyte in the sample
as follows:

       Water

                                 (A,)(I.)
       concentration  (/ig/L) = 	—	
                              (AJ(RF)(V0)
where:
      Ax    =      Area of  characteristic ion  for compound  being
                  measured.
      Is    =      Amount  of internal  standard injected (ng).
      A1S    =      Area of  characteristic  ion  for  the  internal
      	          standard.
      RF    =      Mean relative response factor for compound  being
                  measured  (Sec.  7.2.8).
      V0    =      Volume   of   water   purged   (ml),   taking   into
                  consideration any dilutions made.
                      8240B -  23                         Revision 2
                                                    September 1994

-------
                   Sediment/Soil  Sludge   (on  a  dry-weight  basis)   and  Waste
             (normally on  a wet-weight  basis)

                                              (AJd.)(Vt)
                   concentration  (/^g/kg)  =
                                           (Ais)(RF)(Vi)(Ws)(D)
            where:

                   Ax>  Is> Ais,  RF,  = Same as for water.
                   Vt    =     Volume of total extract (fj.1) (use 10,000 /uL or a
                              factor of this when dilutions are made).

                   V-    =     Volume of extract added (juL) for purging.
                   Ws    =     Weight of sample extracted or purged (g).
                   D    =     % dry weight of sample/100,  or  1 for  a wet-weight
                              basis.

                   7.5.2.3     Where applicable, an estimate of concentration for
            noncalibrated components in the sample should be made.   The formulae
            given  above  should be used with  the  following modifications: The
            areas  Ax and Ais should be from the  total  ion chromatograms, and the
            RF  for the compound  should be assumed  to be 1.  The concentration
            obtained  should be reported  indicating  (1)  that the  value  is an
            estimate  and  (2)  which  internal  standard  was used  to  determine
            concentration.     Use  the  nearest  internal  standard  free  of
            interferences.

                   7.5.2.4     Alternatively, the regression line fitted to the
            initial  calibration  (Sec.  7.2.10.1) may  be used for determination
            of  analyte concentration.


8.0   QUALITY CONTROL

      8.1   Each  laboratory that uses these methods  is  required  to operate a
formal quality control  program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory capability  and  an ongoing analysis of
spiked samples  to evaluate  and  document  data quality.    The  laboratory must
maintain records to document  the  quality  of the data generated.    Ongoing data
quality checks are compared with established performance criteria  to determine
if the results of  analyses meet the performance characteristics of the method.
When results of sample spikes indicate atypical method performance,  a quality
control reference sample must be  analyzed to confirm  that  the measurements were
performed in an in-control mode of operation.

      8.2   Before processing any samples,  the  analyst  should   demonstrate,
through the analysis of  a method blank, that interferences from the analytical
system, glassware, and reagents are under  control.   Each time a set of samples
is extracted or  there is  a change in reagents, a method blank should  be processed
as a  safeguard  against  chronic  laboratory contamination.   The  blank samples
should be carried  through all stages of sample  preparation and measurement.
                                  8240B - 24                        Revision  2
                                                                September  1994

-------
      8.3   The  experience  of  the  analyst  performing  GC/MS  analyses  is
invaluable to the success of the  methods.   Each day that analysis is performed,
the  daily  calibration   standard  should  be  evaluated  to  determine  if  the
chromatographic system is operating properly.  Questions  that should be asked
are:  Do  the  peaks look normal?;  Is  the response obtained comparable  to the
response  from previous  calibrations?   Careful  examination  of the  standard
chromatogram can indicate whether the column is still useable, the injector is
leaking, the injector septum needs replacing, etc.   If any changes are made to
the system (e.g. column changed), recalibration of the system must take place.

      8.4   Required  instrument QC is found  in the following  section:

            8.4.1  The GC/MS system must be tuned  to meet the BFB specifications
      in Sec. 7.2.2.

            8.4.2  There  must be  an  initial  calibration of the GC/MS system as
      specified in Sec. 7.2.

            8.4.3  The GC/MS system must meet the SPCC criteria  specified in Step
      7.3.3 and the CCC criteria in Sec.  7.3.4, each 12 hours.

      8.5   To  establish  the  ability  to  generate  acceptable  accuracy  and
precision, the analyst must perform the following operations.

            8.5.1  A   quality  control   (QC)  reference  sample concentrate  is
      required  containing  each  analyte  at  a  concentration  of  10  mg/L  in
      methanol.  The QC reference sample concentrate may be prepared from pure
      standard materials or purchased as  certified  solutions.  If  prepared by
      the laboratory, the  QC  reference  sample concentrate must  be  made  using
      stock standards prepared independently from those used  for calibration.

            8.5.2  Prepare  a QC  reference  sample to contain  20  /Ltg/L  of each
      analyte by adding 200 /xL of QC reference sample concentrate to 100 mL of
      water.

            8.5.3  Four 5-mL aliquots of the well  mixed QC reference sample are
      analyzed according to the method beginning  in Sec.  7.4.1.

            8.5.4  Calculate the average recovery (x) in jug/L,  and the standard
      deviation of the recovery  (s)  in  jug/L, for each analyte using  the four
      results.

            8.5.5  For each analyte  compare  s  and  x  with the  corresponding
      acceptance criteria_for  precision  and accuracy, respectively,  found in
      Table 6.  If s  and x  for all analytes  meet the acceptance  criteria, the
      system performance  is  acceptable and analysis  of actual samples can_begin.
      If any individual  s exceeds  the  precision limit or any individual x falls
      outside  the  range   for  accuracy,   then  the  system   performance  is
      unacceptable for that analyte.

            NOTE:  The large number of analytes in Table 6 present a substantial
                   probability that one or more will fail at  least  one of the
                                  8240B - 25                        Revision 2
                                                                September 1994

-------
                   acceptance  criteria  when  all  analytes  of a given method are
                   determined.

            8.5.6  When  one  or more of  the analytes tested fail at least one of
      the  acceptance criteria,  the  analyst  must proceed  according  to  Sec.
      8.5.6.1 or 8.5.6.2.

                   8.5.6.1     Locate  and correct the  source of the problem and
            repeat the  test for  all analytes beginning with Sec. 8.5.2.

                   8.5.6.2     Beginning with Sec. 8.5.2,  repeat  the test  only
            for those analytes that failed to meet criteria.  Repeated failure,
            however, will confirm a general problem with the measurement system.
            If this;  occurs,  locate and correct the source  of  the  problem and
            repeat the  test for  all  compounds  of interest beginning with  Sec.
            8.5.2.

      8.6   The laboratory must,  on an  ongoing basis,  analyze a method blank and
a  spiked  replicate   for  each   analytical  batch  (up  to  a  maximum  of  20
samples/batch) to assess accuracy. For soil and waste samples where detectable
amounts of organics are present,  replicate samples may be appropriate in place
of spiked replicates.  For  laboratories analyzing one to ten samples per month,
at least one spiked sample per month  is required.

            8.6.1  The  concentration   of the  spike   in  the  sample  should  be
      determined as follows:

                   8.6.1.1     If, as in compliance monitoring, the concentration
            of a  specific  analyte in  the  sample is  being checked  against  a
            regulatory  concentration limit, the  spike  should  be  at that limit
            or 1 to 5 times higher than the  background concentration determined
            in Sec. 8.6.2, whichever concentration would  be larger.

                   8.6.1.2     If the  concentration of  a  specific  analyte  in  a
            water  sample is not  being checked against a  specific  limit,  the
            spike  should  be  at  20  jug/L  or  1  to  5  times  higher than  the
            background   concentration  determined  in  Sec.  8.6.2,  whichever
            concentration  would  be larger.   For other  matrices,  recommended
            spiking  concentration is 10 times  the EQL.

            8.6.2  Analyze  one  5-mL sample aliquot to determine the background
      concentration  (B) of  each analyte.   If  necessary, prepare  a  new  QC
      reference sample concentrate (Sec. 8.5.1) appropriate for the background
      concentration  in  the sample.   Spike  a  second  5-mL  sample  aliquot  with
      10 fj.L of the QC reference  sample concentrate and analyze it to determine
      the  concentration after spiking  (A)  of each  analyte.   Calculate  each
      percent recovery  (p)  as  100(A-B)%/T,  where  T is  the known  true value of
      the  spike.

            8.6.3  Compare  the percent  recovery  (p) for each analyte in a water
      sample with  the  corresponding  QC acceptance criteria found  in Table 6.
      These acceptance  criteria were  calculated  to  include an  allowance for
      error in measurement of both  the background  and  spike  concentrations,


                                  8240B - 26                        Revision 2
                                                                September 1994

-------
      assuming a spike to background  ratio of 5:1.  This error will be accounted
      for to the extent that the analyst's spike  to background ratio approaches
      5:1.  If spiking was performed  at  a concentration  lower than 20 M9/U the
      analyst must use either the QC acceptance  criteria presented in Table 6,
      or  optional  QC  acceptance criteria calculated  for the  specific spike
      concentration.  To calculate optional acceptance criteria for the recovery
      of  an  analyte:  (1) Calculate accuracy (x')  using the  equation found in
      Table  7,  substituting the spike concentration  (T)  for C; (2)  calculate
      overall precision  (S') using the equation in  Table 7, substituting x' for
      x;  (3)  calculate the  range  for  recovery at the  spike concentration as
      (100x'/T) + 2.44(100S7T)%.

            8.6.4  If  any individual  p falls outside  the designated  range for
      recovery,  that  analyte  has  failed the  acceptance  criteria.    A check
      standard containing each analyte that failed the criteria must be analyzed
      as described in Sec.  8.7.

      8.7    If any analyte  in a  water sample fails the  acceptance criteria for
recovery in Sec.  8.6,  a QC reference  sample containing each analyte that failed
must be prepared and analyzed.

      NOTE: The  frequency for the required analysis  of a  QC reference sample
            will depend upon the number of analytes being simultaneously tested,
            the  complexity of the sample matrix,  and  the performance  of the
            laboratory.   If  the entire  list  of  analytes  in Table  6  must be
            measured  in  the sample  in  Sec.  8.6,  the probability  that  the
            analysis of a QC reference sample will be required is high.   In this
            case, the QC reference sample  should  be routinely  analyzed with the
            spiked sample.

            8.7.1  Prepare the QC reference sample by adding 10 pi  of the QC
      reference  sample  concentrate (Sec.  8.5.1  or  8.6.2)  to 5  ml of  reagent
      water.  The QC  reference  sample  needs  only to  contain  the analytes that
      failed criteria in  the  test in Sec.  8.6.

            8.7.2 Analyze the QC reference sample to determine the concentration
      measured (A) of each analyte.   Calculate  each  percent recovery  (ps) as
      100(A/T)%, where T  is  the  true value of  the standard concentration.

            8.7.3 Compare the percent  recovery (ps)  for each analyte with the
      corresponding QC acceptance criteria found  in Table 6.   Only analytes that
      failed the test in Sec. 8.6 need to be compared with these criteria.  If
      the recovery of any such analyte falls outside the designated range, the
      laboratory performance  for that analyte  is judged to  be out  of control,
      and the problem must be immediately  identified and corrected.  The result
      for that analyte in the unspiked sample is suspect and may  not be reported
      for regulatory compliance  purposes.

      8.8   As part of  the  QC program  for the  laboratory,  method accuracy for
each matrix studied must be assessed  and records  must  be maintained.  After the
analysis of five spiked samples _(of the same matrix)  as in Sec. 8.6,  calculate
the  average  percent  recovery  (p)  and  the standard  deviation  of  the  percent
recovery (sp).   Express  the accuracy  assessment as  a  percent  recovery interval


                                  8240B -  27                        Revision 2
                                                                September 1994

-------
from p - 2sp to p +  2sp.   If p = 90% and sp = 10%,  for  example,  the accuracy
interval  is  expressed as  70-110%.   Update  the  accuracy assessment  for  each
analyte  on  a  regular  basis  (e.g.,  after  each  five  to  ten  new  accuracy
measurements).

      8.9   To determine acceptable accuracy and precision limits for surrogate
standards the following procedure should be performed.

            8.9.1 For each sample  analyzed,  calculate the  percent recovery of
      each surrogate  in the  sample.

            8.9.2 Once a minimum of thirty samples of the same matrix have been
      analyzed,  calculate  the  average  percent   recovery  (P)  and  standard
      deviation of the percent recovery (s) for each of the surrogates.

            8.9.3 For a  given matrix,  calculate  the upper and  lower control
      limit for method performance  for each  surrogate  standard.  This should be
      done as follows:

            Upper Control  Limit  (UCL) = P + 3s
            Lower Control  Limit  (LCL) = P - 3s

            8.9.4 For aqueous  and  soil  matrices,  these laboratory established
      surrogate  control  limits  should,  if applicable,  be  compared  with  the
      control limits listed in Table 8.  The  limits given in Table 8 are multi-
      laboratory performance based limits  for  soil  and aqueous  samples,  and
      therefore, the  single-laboratory  limits  established in  Sec.  8.9.3  must
      fall  within those given  in Table 8 for these matrices.

            8.9.5 If  recovery is not within limits, the following procedures are
      required.

            •     Check  to  be  sure  there  are  no   errors  in  calculations,
                  surrogate  solutions  and  internal   standards.   Also,  check
                  instrument performance.

            •     Recalculate  the  data  and/or  reanalyze  the  extract if any of
                  the above  checks  reveal a  problem.

            •     Reextract  and  reanalyze  the  sample if  none  of the above are
                  a  problem  or flag the data as "estimated concentration".

            8.9.6 At  a minimum, each laboratory should update surrogate recovery
      limits on a matrix-by-matrix  basis, annually.

      8.10   It  is  recommended  that the  laboratory  adopt additional  quality
assurance practices for use with this method.  The specific practices that are
most productive depend upon  the  needs of  the laboratory  and  the nature of the
samples.    Field duplicates  may be analyzed  to  assess  the  precision  of  the
environmental measurements.  When doubt exists over the identification of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar column or a different ionization  mode using a mass spectrometer must


                                  8240B - 28                        Revision 2
                                                                September 1994

-------
be used.  Whenever  possible,  the laboratory should analyze standard reference
materials and participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1   This method was tested by 15 laboratories using organic-free reagent
water, drinking water, surface water, and industrial wastewaters spiked at six
concentrations over the range  5-600 jug/L.   Single operator precision, overall
precision,  and method  accuracy  were  found  to  be directly  related to  the
concentration of the analyte and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 7.


10.0  REFERENCES

1.    U.S. EPA 40  CFR  Part 136, "Guidelines Establishing Test Procedures for the
      Analysis  of  Pollutants   Under  the   Clean  Water  Act,  Method  624,"
      October 26, 1984.

2.    U.S.  EPA Contract  Laboratory Program,  Statement  of Work  for  Organic
      Analysis, July  1985, Revision.

3.    Bellar, T.A., and J.J. Lichtenberg, J. Amer. Water Works Assoc., 66(12),
      739-744, 1974.

4.    Bellar, T.A.,  and J.J.  Lichtenberg,  "Semi-Automated  Headspace Analysis of
      Drinking  Waters and  Industrial  Waters  for Purgeable  Volatile  Organic
      Compounds," in Van Hall, ed., Measurement of Organic  Pollutants in Water
      and Wastewater, ASTM STP 686, pp. 108-129,  1979.

5.    Budde, W.L. and J.W. Eichelberger,  "Performance Tests for the Evaluation
      of  Computerized  Gas  Chromatography/Mass  Spectrometry  Equipment  and
      Laboratories,"  EPA-600/4-79-020,  U.S. Environmental  Protection  Agency,
      Environmental Monitoring and Support Laboratory, Cincinnati, Ohio 45268,
      April 1980.

6.    Eichelberger, J.W., L.E. Harris, and  W.L.  Budde,  "Reference Compound to
      Calibrate   Ion   Abundance  Measurement   in  Gas   Chromatography-Mass
      Spectrometry Systems," Analytical Chemistry, 47, 995-1000, 1975.

7.    "Method Detection Limit for Methods  624 and  625," Olynyk, P., W.L. Budde,
      and J.W. Eichelberger, Unpublished report, October  1980.

8.    "Interlaboratory Method Study for EPA Method 624-Purgeables,"  Final Report
      for EPA Contract 68-03-3102.

9.    "Method Performance Data for Method 624,"  Memorandum from R. Slater and
      T.  Pressley,   U.S.   Environmental    Protection   Agency,  Environmental
      Monitoring and  Support  Laboratory,  Cincinnati, Ohio  45268,  January 17,
      1984.
                                  8240B - 29                        Revision 2
                                                                September 1994

-------
10.    Gebhart, J.E.;  Lucas,  S.V.;  Naber,  S.J.;  Berry,  A.M.; Danison,  T.H.;
      Burkholder, H.M.  "Validation of SW-846 Methods 8010, 8015, and 8020"; U.S.
      Environmental  Protection  Agency,  Environmental  Monitoring and  Support
      Laboratory, Cincinnati, Old 45268, July 1987, Contract No.  68-03-1760.

11.    Lucas, S.V.; Kornfeld, R.A. "GC-MS Suitability  Testing  of  RCRA Appendix
      VIII and Michigan List Analytes "; U.S. Environmental Protection Agency,
      Environmental Monitoring  and  Support Laboratory, Cincinnati,  OH 45268,
      February 20, 1987, Contract No. 68-03-3224.
                                  8240B - 30                        Revision 2
                                                                September 1994

-------
                                   TABLE  1.
        RETENTION TIMES AND CHARACTERISTIC IONS FOR VOLATILE COMPOUNDS
Compound
Retention
Time (minutes)
Primary Ion  Secondary Ion(s)
Ethylene oxide
Chloromethane
Di chl orodi f 1 uoromethane
Bromomethane
Vinyl chloride
Acetonitrile
Chloroethane
Methyl iodide
Methylene chloride
Carbon disulfide
Tri chl orof 1 uoromethane
Propionitrile
Ally! chloride
1,1-Dichloroethene
Bromochloromethane (I.S.)
Ally! alcohol
trans-l,2-Dichloroethene
1,2-Dichloroethane
Propargyl alcohol
Chloroform
l,2-Dichloroethane-d4(surr)
2-Butanone
Methacrylonitrile
Dibromomethane
2-Chloroethanol
b-Propiolactone
Epichlorohydrin
1,1,1 -Tri chloroethane
Carbon tetrachlon'de
1,4-Dioxane
Isobutyl alcohol
Bromod i chl oromethane
Chloroprene
l,2:3,4-Diepoxybutane
1,2-Dichloropropane
Chloral hydrate (b)
cis-l,3-Dichloropropene
Bromoacetone
Trichloroethene
Benzene
trans-l,3-Dichloropropene
1 , 1 , 2-Tri chl oroethane
3-Chloropropionitrile
1,2-Dibromoethane
Pyridine
1.30
2.30
2.47
3.10
3.80
3.97
4.60
5.37
6.40
7.47
8.30
8.53
8.83
9.00
9.30
9.77
10.00
10.10
10.77
11.40
12.10
12.20
12.37
12.53
12.93
13.00
13.10
13.40
13.70
13.70
13.80
14.30
14.77
14.87
15.70
15.77
15.90
16.33
16.50
17.00
17.20
17.20
17.37
18.40
18.57
44
50
85
94
62
41
64
142
84
76
101
54
76
96
128
57
96
62
55
83
65
72
41
93
49
42
57
97
117
88
43
83
53
55
63
82
75
136
130
78
75
97
54
107
79
44, 43, 42
52, 49
85, 87, 101, 103
96, 79
64, 61
41, 40, 39
66, 49
142, 127, 141
49, 51, 86
76, 78, 44
103, 66
54, 52, 55, 40
76, 41, 39, 78
61, 98
49, 130, 51
57, 58, 39
61, 98
64, 98
55, 39, 38, 53
85, 47
102
43, 72
41, 67, 39, 52, 66
93, 174, 95, 172, 176
49, 44, 43, 51, 80
42, 43, 44
57, 49, 62, 51
99, 117
119, 121
88, 58, 43, 57
43, 41, 42, 74
85, 129
53, 88, 90, 51
55, 57, 56
62, 41
44, 84, 86, 111
77, 39
43, 136, 138, 93, 95
95, 97, 132
52, 71
77, 39
83, 85, 99
54, 49, 89, 91
107, 109, 93, 188
79, 52, 51, 50
                                  8240B -  31
                                       Revision 2
                                   September  1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (minutes)
Primary Ion  Secondary Ion(s)
2-Chloroethyl vinyl ether
2-Hydroxypropionitrile
1,4-Difluorobenzene (I.S.)
Malononitrile
Methyl methacryl ate
Bromoform
1,1,1 , 2-Tetrachl oroethane
l,3-Dichloro-2-propanol
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
1 , 2 , 3-Tr i chl oropropane
l,4-Dichloro-2-butene
n-Propylamine
2-Picoline
Toluene
Ethyl methacryl ate
Chlorobenzene
Pentachl oroethane8
Ethyl benzene
1, 2 -Dibromo-3- chl oropropane
4-Bromofluorobenzene (surr.)
Benzyl chloride
Styrene
bis-(2-Chloroethyl) sulfide(b)
Acetone
Acrolein
Acrylonitrile
Chlorobenzene-d5 (I.S.)
Chlorodibromomethane
1,1-Dichloroethane
Ethanol
2-Hexanone
lodomethane
4-Methyl -2-pentanone
Toluene-d8 (surr.)
Vinyl acetate
Xylene (Total)
18.60
18.97
19.60
19.60
19.77
19.80
20.33
21.83
22.10
22.20
22.20
22.73
23.00
23.20
23.50
23.53
24.60
24.83
26.40
27.23
28.30
29.50
30.83
33.53
--
--
--
--
--
--
--
--
--
--
--
--
_ _
63
44
114
66
69
173
131
79
83
164
75
75
59
93
92
69
112
167
106
157
95
91
104
109
43
56
53
117
129
63
31
43
142
43
98
43
106
65,106
44,43,42,53
63,88
66,39,65,38
69,41,100,39
171,175,252
131,133,117,119,95
79,43,81,49
85,131,133
129,131,166
75,77,110,112,97
75,53,77,124,89
59,41,39
93,66,92,78
91,65
69,41,99,86,114
114,77
167,130,132,165,169
91
157,75,155,77
174,176
91,126,65,128
104,103,78,51,77
111, 158, 160
58
55,58
52,51
82,119
208,206
65,83
45,27,46
58,57, 100
127,141
58,57,100
70,100
86
91
a The base peak at m/e 117 was  not used due to an interference at  that mass with
  a nearly coeluting  internal  standard,  chlorobenzene-d5.
b  Response factor judged  to be too low  (less than 0.02) for practical use.
(I.S.) = Internal Standard
(surr) = Surrogate
                                  8240B - 32
                                        Revision 2
                                    September 1994

-------
                        TABLE  2.
ESTIMATED QUANTITATION LIMITS (EQL) FOR VOLATILE ORGANICS
                           Estimated
                          Quantitation
                            Limits"
Ground water
Volatiles /j.g/1
Acetone
Acetonitrile
Ally! chloride
Benzene
Benzyl chloride
Bromodi chl oromethane
Bromoform
Bromomethane
2-Butanone
Carbon disulfide
Carbon tetrachloride
Chlorobenzene
Chlorodibromomethane
Chloroethane
2-Chloroethyl vinyl ether
Chloroform
Chl oromethane
Chloroprene
1 , 2-Di bromo-3-chl oropropane
1,2-Dibromoethane
Dibromomethane
l,4-Dichloro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1 Dichloroethene
trans- 1, 2-Di chl oroethene
1, 2-Di chl oropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
Ethyl benzene
Ethyl methacrylate
2-Hexanone
Isobutyl alcohol
Methacrylonitrile
Methyl ene chloride
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Pentachloroethane
100
100
5
5
100
5
5
10
100
100
5
5
5
10
10
5
10
5
100
5
5
100
5
5
5
5
5
5
5
5
5
5
50
100
100
5
5
5
50
10
Low Soil/Sediment"
M9/kg
100
100
5
5
100
5
5
10
100
100
5
5
5
10
10
5
10
5
100
5
5
100
5
5
5
5
5
5
5
5
5
5
50
100
100
5
5
50
50
10
                       8240B - 33                        Revision 2
                                                     September 1994

-------
                                   TABLE 2.
                                  (Continued)
                                      Estimated
                                     Quantitation
                                       Limits8
                            Ground water        Low  Soil/Sedimentb
Volatiles                       p.g/1
Propionitrile
Styrene
1,1,1, 2 -Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachl oroethene
Toluene
1,1,1-Trichl oroethane
1 , 1 , 2-Trichl oroethane
Trichloroethene
1,2,3-Trichloropropane
Vinyl acetate
Vinyl chloride
Xylene (Total)
100
5
5
5
5
5
5
5
5
5
50
10
5
100
5
5
5
5
5
5
5
5
5
50
10
5
a Sample EQLs are highly matrix dependent.   The  EQLs  listed herein are provided
  for guidance and may not always be achievable.

b EQLs  listed  for soil/sediment are based  on wet weight.   Normally data are
  reported on a dry weight basis; therefore,  EQLs will be higher, based on the
  percent dry weight of each  sample.
               Other Matrices                      Factor0
               Water miscible liquid waste             50
               High-concentration soil and sludge     125
               Non-water miscible waste               500


  CEQL =  [EQL for  low soil/sediment  (see Table  2)] X  [Factor  found in this
          table].  For non-aqueous samples, the factor is on a wet weight basis.
                                  8240B - 34                        Revision 2
                                                                September  1994

-------
                                TABLE 3.
                     BFB KEY ION ABUNDANCE CRITERIA
    Mass              Ion Abundance Criteria .
    50                15 to 40% of mass 95
    75                30 to 60% of mass 95
    95                base peak, 100% relative abundance
    96                5 to 9% of mass 95
   173                less than 2% of mass 174
   174                greater than 50% of mass 95
   175                5 to 9% of mass 174
   176                greater than 95% but less than 101% of mass 174
   177                5 to 9% of mass 176
                                TABLE 4.
           QUANTITY  OF  METHANOL  EXTRACT  REQUIRED  FOR ANALYSIS
                  OF HIGH-CONCENTRATION  SOILS/SEDIMENTS
       Approximate                               Volume of
   Concentration Range                        Methanol  Extract
a
      500- 10,000 Mg/kg                            100 /xL
    1,000- 20,000 jugAg                             50 ML
    5,000-100,000 jug/kg                             10 ML
   25,000-500,000 Mg/kg                            100 ML of 1/50 dilution"
Calculate  appropriate  dilution  factor  for  concentrations  exceeding  this
table.

a  The volume of methanol  added  to  5 mL of water being purged should be kept
   constant. Therefore, add to the 5 mL  syringe whatever volume of methanol
   is necessary to maintain a volume of  100 juL added to the syringe.

b  Dilute  and  aliquot  of  the  methanol  extract  and  then take  100 ML  for
   analysis.


                               8240B - 35                         Revision  2
                                                             September 1994

-------
                                   TABLE 5.
       VOLATILE INTERNAL STANDARDS WITH CORRESPONDING ANALYTES ASSIGNED
                               FOR QUANTITATION
Bromochloromethane

Acetone
Acrolein
Acrylonitrile
Bromomethane
Carbon disulfide
Chloroethane
Chloroform
Chioromethane
Di chlorodi f1uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
l,2-Dichloroethane-d4 (surrogate)
1,1-Dichloroethene
trans-l,2-Dichloroethene
lodomethane
Methylene chloride
Tri chlorof1uoromethane
Vinyl chloride
1,4-Di f1uorobenzene

Benzene
Bromodichloromethane
Bromoform
2-Butanone
Carbon tetrachloride
Chlorodi bromomethane
2-Chloroethyl vinyl ether
Dibromomethane
l,4-Dichloro-2-butene
1,2-Dichloropropane
cis-1,3-Dichloropropene
trans-l,3-Dichloropropene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Vinyl acetate
                        Chlorobenzene-dr
                        Bromofluorobenzene  (surrogate)
                        Chlorobenzene
                        Ethyl benzene
                        Ethyl  methacrylate
                        2-Hexanone
                        4-Methyl-2-pentanone
                        Styrene
                        1,1,2,2-Tetrachloroethane
                        Tetrachloroethene
                        Toluene
                        Toluene-d8  (surrogate)
                        1,2,3-Trichloropropane
                        Xylene
                                  8240B - 36
                      Revision 2
                  September 1994

-------
                                TABLE 6.
                 CALIBRATION AND QC  ACCEPTANCE  CRITERIA8


Parameter
Benzene
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
2-Chloroethyl vinyl ether
Chloroform
Chl oromethane
Di bromochl oromethane
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans -1, 2-Di chl oroethene
1,2-Dichloropropane
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
Ethyl benzene
Methyl ene chloride
1,1, 2, 2-Tetrachl oroethane
Tetrachl oroethene
Toluene
1 , 1 , 1 -Tri chl oroethane
1 , 1 , 2-Trichl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
Vinyl chloride
Range
for Q
(M9/L)
12.8-27.2
13.1-26.9
14.2-25.8
2.8-37.2
14.6-25.4
13.2-26.8
D-44.8
13.5-26.5
D-40.8
13.5-26.5
12.6-27.4
14.6-25.4
12.6-27.4
14.5-25.5
13.6-26.4
10.1-29.9
13.9-26.1
6.8-33.2
4.8-35.2
10.0-30.0
11.8-28.2
12.1-27.9
12.1-27.9
14.7-25.3
14.9-25.1
15.0-25.0
14.2-25.8
13.3-26.7
9.6-30.4
0.8-39.2
Q = Concentration measured in QC check
Limit
for s
(M9/L)
6.9
6.4
5.4
17.9
5.2
6.3
25.9
6.1
19.8
6.1
7.1
5.5
7.1
5.1
6.0
9.1
5.7
13.8
15.8
10.4
7.5
7.4
7.4
5.0
4.8
4.6
5.5
6.6
10.0
20.0
sample,
Range
for x
(MgA)
15.2-26.0
10.1-28.0
11.4-31.1
D-41.2
17.2-23.5
16.4-27.4
D-50.4
13.7-24.2
D-45.9
13.8-26.6
11.8-34.7
17.0-28.8
11.8-34.7
14.2-28.4
14.3-27.4
3.7-42.3
13.6-28.4
3.8-36.2
1.0-39.0
7.6-32.4
17.4-26.7
D-41.0
13.5-27.2
17.0-26.6
16.6-26.7
13.7-30.1
14.3-27.1
18.5-27.6
8.9-31.5
D-43.5
in jug/L.
Range
P.Ps
(%)
37-151
35-155
45-169
D-242
70-140
37-160
D-305
51-138
D-273
53-149
18-190
59-156
18-190
59-155
49-155
D-234
54-156
D-210
D-227
17-183
37-162
D-221
46-157
64-148
47-150
52-162
52-150
71-157
17-181
D-251

s_ = Standard deviation of four recovery measurements, in p,g/L.
x = Average recovery
p, ps = Percent recovery
D = Detected; result
for four recovery
measured.
measurements, in /ng/L.



must be greater than zero.
Criteria from 40 CFR Part 136 for Method 624 and were calculated assuming a
QC check sample concentration of  20 ng/l.  These criteria are based directly
upon the method performance data  in Table 7.  Where necessary, the limits for
recovery  have  been  broadened to  assure applicability  of  the limits  to
concentrations below those used to develop Table 7.
                               8240B - 37
    Revision 2
September 1994

-------
                                   TABLE 7.
         METHOD ACCURACY AND PRECISION AS  FUNCTIONS OF CONCENTRATION8
Parameter
Benzene
Bromodi chl oromethane
Bromoform
Bromomethane
Carbon tetrachloride
Chlorobenzene
Chloroethane
2-Chloroethyl vinyl ether8
Chloroform
Chl oromethane
Di bromochl oromethane
1 , 2 -Di chl orobenzeneb
1,3-Dichlorobenzene
l,4-Dichlorobenzeneb
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans - 1 , 2 , -Di chl oroethene
1 , 2-Di chl oropropane8
cis-1 ,3-Dichloropropenea
trans- 1,3-Di chl oropropene"
Ethyl benzene
Methyl ene chloride
1,1,2,2-Tetrachloroethane
Tetrachl oroethene
Toluene
1,1,1-Tri chloroethane
1,1, 2 -Tri chloroethane
Trichloroethene
Trichlorofluoromethane
Vinyl chloride
Accuracy, as
recovery, x'
(M9/L)
0.93C+2.00
1.03C-1.58
1.18C-2.35
l.OOC
1.10C-1.68
0.98C+2.28
1.18C+0.81
l.OOC
0.93C+0.33
1.03C-1.81
1.01C-0.03
0.94C+4.47
1.06C+1.68
0.94C+4.47
1.05C+0.36
1.02C+0.45
1.12C+0.61
1.05C+0.03
l.OOC
l.OOC
l.OOC
0.98C+2.48
0.87C+1.88
0.93C+1.76
1.06C+0.60
0.98C+2.03
1.06C+0.73
0.95C+1.71
1.04C+2.27
0.99C+0.39
l.OOC
Single analyst
precision, sr'
(M9/L)
0.26X-1.74
0.15X+0.59
0.12X+0.34
0.43x
0.12X+0.25
O.lSx-0.09
O.Hx+2.78
0.62x
0.16x+0.22
0.37X+2.14
0.17X-0.18
0.22X-1.45
O.Hx-0.48
0.22X-1.45
0.13X-0.05
0.17x-0.32
0.17X+1.06
0.14X+0.09
0.33x
0.38x
0.25x
0.14X+1.00
0.15X+1.07
0.16X+0.69
0.13X-0.18
0.15X-0.71
0.12X-0.15
0.14x+0.02
O.lSx+0.36
0.33X-1.48
0.48x
Overall
precision,
S' (/*g/L)
0.25x-1.33
0.20x+1.13
0.17X+1.38
0.58x
O.llx+0.37
0.26x-1.92
0.29X+1.75
0.84x
O.lSx+0.16
0.58x+0.43
0.17X+0.49
0.30X-1.20
O.lSx-0.82
0.30x-1.20
0.16X+0.47
0.21X-0.38
0.43X-0.22
0.19X+0.17
0.45x
0.52x
0.34X
0.26X-1.72
0.32X+4.00
0.20X+0.41
0.16X-0.45
0.22x-1.71
0.21X-0.39
O.lSx+0.00
0.12X+0.59
0.34x-0.39
0.65x
x'    =      Expected  recovery  for  one  or  more  measurements   of  a  sample
             containing  a  concentration  of C,  in jug/L.
sr'    =      Expected  single analyst standard deviation  of  measurements at an
             average concentration  of x,  in  jug/L-
S'    =      Expected  interlaboratory  standard deviation of measurements at an
             average concentration  found of  x,  in  /xg/L.
C     =      True  value  for  the  concentration,  in  M9/L-
x     =      Average  recovery found for measurements of  samples  containing a
             concentration of C,  in fj.g/1.
a     Estimates based upon  the  performance  in  a single  laboratory.
b     Due  to chromatographic resolution problems,  performance statements  for
      these  isomers are based upon the  sums of their concentrations.
                                  8240B - 38
    Revision 2
September 1994

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR WATER AND SOIL/SEDIMENT SAMPLES
                                  Low/High             Low/High
Surrogate Compound                Water             Soil/Sediment
4-Bromofluorobenzene             86-115               74-121
l,2-Dichloroethane-d4            76-114               70-121
Toluene-dQ                       88-110               81-117
                                  8240B -  39                        Revision 2
                                                                September 1994

-------
                          FIGURE 1.
                      PURGING  CHAMBER
FOAMTN*
                       GOT M W. OJ)
                       14 MM O.O.
                       INIET 1M IN. O.O.
EXT 1M M. O.O
10 MM OLA88
MEDIUM HJKJtirt
SAMPLE INIXT

»WAY SYNNOE VALVC

17 CM aO OAUOE SVWNQE NEEOU

• MM O.O flUeSER SEPTUM

INlfT 1M IN O.O.
                                                         o o
                                                    13X
                                                    MOt^CUiAR SICVC
                                                    PU*WE GAS FUTCT
                                                     PUMQEQAS-
                                                     FLOWCONTWX
                          8240B  -  40
                                     Revision  2
                                September 1994

-------
                FIGURE  2.
TRAP PACKINGS AND CONSTRUCTION TO INCLUDE
    DESORB  CAPABILITY  FOR  METHOD  8240B
                       CONSTRUCTION OCTML
                8240B - 41
    Revision 2
September 1994

-------
                             FIGURE 3.
  SCHEMATIC OF  PURGE-AND-TRAP DEVICE - PURGE MODE FOR METHOD  8240B
GAMIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE QAS
PLOW CONTROL
13X MOLECULAR
SIEVE FILTER
LIQUID INJECTION PORTS

   r— COLUMN OVEN
                                    LJUUVr
                                               CONFIRMATORY COLUMN
                                              TO DETECTOR
                                               - ANALYTICAL COLUMN
                              OPTIONAL 4PORT COLUMN
                              SELECTION VALVE
                                       TRAP INLET
                                     TRAP
                                     2TC
                                PURGING
                                DEVICE
             NOTE
             ALL LINES BETWEEN TRAP
             AMD QC SHOULD BE HEATED
             TOID-C.
                             8240B - 42
                              Revision 2
                          September 1994

-------
                             FIGURE  4.
SCHEMATIC OF PURGE-AND-TRAP  DEVICE  -  DESORB MODE FOR METHOD 8240B
 CARRKRGAS
 FLOW CONTROL
 PRESSURE
 REGULATOR
UOWO INJECTION PORTS

    -COLUMN OVEN
 PURGE GAS
 FLOW CONTROL
 13X MOLECULAR
 SIEVE FILTER
               CONFIRMATORY COLUMN


              TO DETECTOR

               ANALYTICAL COLUMN
                               OPTIONAL tPORT COLUMN
                               SELECTION VALVE
                                        TRAP INLET
                                       TRAP
                                       2QO*C
                                  PURQMG
                                  DEVICE
              NOTE
              ALL LINES BETWEEN TRAP
              AND GC SHOULD BE HEATED
              TO WC.
                             8240B - 43
                             Revision 2
                         September 1994

-------
                    FIGURE 5.
                LOW SOILS IMPINGER
  PURGE INLET FITTING
 SAMPLE OUTLET FITTING
9" i 6mm 0 0 CLASS TUBING f fc-
                                     SEPTUM
                                        CAP
            40ml VIAL
                   8240B  - 44
    Revision 2
September 1994

-------
                                 METHOD 8240B
VOLATILE ORGANICS  BY  GAS  CHROMATOGRAPHY/MASS  SPECTROMETRY  (GC/MS)
                                              Water
                                             Miscible
                                              Liquids
                                                                        Soil/Sediment
                                                                         and Waste
                                                                          Samples
    7.1
   Select
procedure for
 introducing
 sample into
  GC/MS.
                              7.4.2.1
                            Dilute sample
                             at least 50
                              fold with
                               water.
                                                                                        7.4.3 Screen
                                                                                        sample using
                                                                                        Method 3810
                                                                                          or 3820.
 Direct
Injection
                                                                  screening
                                                                method for the
                                                                    waste
                                                                   matrix.
                                                   Water
                                                   Samples
rurge-and-trap
                                                        7.4.1.1
                                                     Screen sample
                                                     using Method
                                                     3810 or 3820.
            7.2.1
          Set GC/MS
           operating
          conditions.
                                                        7.4.1.7
                                                        Perform
                                                       secondary
                                                       dilutions.
        7.2.4 Connect
        purge-and-trap
        device to GC.
 7.2.6 Perform
 purge-and-trap
    analysis.
                                                              7.4.1.8 Add
                                                            internal standard
                                                             and surrogate
                                                            spiking solutions
     7.2.8
 Calculate RFs
 for 5 SPCCs.
                                                               7.4.1.10
                                                                Perform
                                                             purge-and-trap
                                                               procedure.
  7.3 Perform
     daily
   calibration
  using SPCCs
   and CCCs.
                                   8240B -  45
                                                                                     Revision 2
                                                                                September 1994

-------
                                      METHOD  8240B
                                       (continued)
    7.4.3
      Is
 concentration
  > 1 mg/Kg?
   7.4.3.1.1
 Choose sample
 size based on
   estimated
 concentration.
 7.4.3.1.3 Add
internal standard
 and surrogate
spiking solutions.
   7.4.3.1.5
   Determine
   percent dry
   weight of
    sample.
    7.4.3.1.7
     Perform
  purge-and-trap
    procedure.
7.4.3.2 Choose
  solvent for
 extraction or
dilution. Weigh
    sample.
 7.4.3.2.2 Add
    solvent,
    shake.
   7.4.3.2.7
    Perform
 purge-and-trap
  procedure.
    7.4.1.11
   Attach trap
   to GC and
    perform
    analysis.
7.5.1.1  Indentify
   analytes by
 comparing the
sample retention
time and sample
 mass spectra.
7.5.2.2 Calculate
the concentration
of each identified
    analyte.
                                            7.5.2.4
                                           Report all
                                            results.
                                         f   Stop    J
                                        8240B  -  46
                                                 Revision  2
                                            September 1994

-------
                                 METHOD 8250A

SEMIVOLATILE ORGANIC COMPOUNDS BV GAS CHROMAT06RAPHY/MASS SPECTROMETRY (GC/MS)


1.0   SCOPE AND APPLICATION

      1.1   Method 8250 is used to determine the concentration of semivolatile
organic compounds in extracts prepared from all types of solid waste matrices,
soils, and ground water.  Direct  injection  of  a  sample  may be used in limited
applications.  The following compounds can be determined by this method:
                                            Appropriate Preparation Techniques
Compounds
CAS Noa    3510
                                  8250A - 1
3520  3540/   3550  3580
      3541
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
Aldrin
4-Aminobiphenyl
Aniline
Anthracene
Aroclor - 1016 (PCB-1016)
Aroclor - 1221 (PCB-1221)
Aroclor - 1232 (PCB-1232)
Aroclor - 1242 (PCB-1242)
Aroclor - 1248 (PCB-1248)
Aroclor - 1254 (PCB-1254)
Aroclor - 1260 (PCB-1260)
Benzidine
Benzoic acid
Benz(a)anthracene
Benzo(b)f 1 uoranthene
Benzo(k)fluoranthene
Benzo(g,h,i)perylene
Benzo(a)pyrene
Benzyl alcohol
a-BHC
iS-BHC
S-BHC
7-BHC (Lindane)
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
83-32-9

208-96-8
98-86-2
309-00-2
92-67-1
62-53-3
120-12-7
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
92-87-5
65-85-0
56-55-3
205-99-2
207-08-9
191-24-2
50-32-8
100-51-6
319-84-6
319-85-7
319-86-8
58-89-9
111-91-1
111-44-4
108-60-1
117-81-7
101-55-3
85-68-7
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
ND
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
X
X
X
X
X
X
X
X
CP
ND
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
ND
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
                                Revision 1
                            September 1994

-------
Compounds
        Appropriate Preparation Techniques

CAS No"   3510     3520  3540/   3550  3580
                        3541
Chlordane (technical)
4-Chloroaniline
1 -Chi oronaphthal ene
2-Chloronaphthalene
4-Chl oro- 3 -methyl phenol
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Chrysene-d12 (I.S.)
4,4'-DDD
4,4'-DDT
4, 4' -DDE
Dibenz(a,j)acridine
Di benz ( a , h ) anthracene
Dibenzofuran
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S)
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dieldrin
Diethyl phthalate
Di methyl ami noazobenzene
7,12-Dimethylbenz(a)-
anthracene
a , o-Dimethyl phenethyl ami ne
2,4-Dimethylphenol
Dimethyl phthalate
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di phenyl ami ne
1, 2 -Di phenyl hydrazine
Di-n-octyl phthalate
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
Ethyl methanesulfonate
57-74-9
106-47-8
90-13-1
91-58-7
59-50-7
95-57-8
7005-72-3
218-01-9

72-54-8
50-29-3
72-55-9
224-42-0
53-70-3
132-64-9
84-74-2
95-50-1
541-73-1
106-46-7
3855-82-1
91-94-1
120-83-2
87-65-0
60-57-1
84-66-2
60-11-7

57-97-6
122-09-8
105-67-9
131-11-3
534-52-1
51-28-5
121-14-2
606-20-2
122-39-4
122-66-7
117-84-0
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
62-50-0
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

CP(45)
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X

X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
ND
X
X
ND

ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
ND
X
X

X
X
X
X
X
X
X
ND
X
ND
X
X
X
X
X
X
X
ND
X
X
ND

ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
X
ND
X
X

X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
ND
X
X
ND

ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
                                   8250A - 2
                                Revision 1
                            September 1994

-------
Compounds
        Appropriate Preparation Techniques

CAS Noa   3510     3520  3540/   3550  3580
                        3541
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachlorocyclopentadiene
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Methoxychlor
3-Methyl chol anthrene
Methyl methanesulfonate
2-Methylnaphthalene
2-Methyl phenol
4-Methylphenol
Naphthalene
Naphthalene-d8 (I.S.)
1-Naphthylamine
2-Naphthylamine
2-Nitroanil ine
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
Nitrobenzene-d5 (surr.)
2-Nitrophenol
4-Nitrophenol
N-Nitrosodi butyl ami ne
N-Nitrosodi methyl ami ne
N-Nitrosodiphenylamine
N-Nitrosodi-n-propylamine
N-Nitrosopi peri dine
Pentachl orobenzene
Pentachloronitrobenzene
Pentachlorophenol
Perylene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenol
Phenol -d6 (surr.)
2-Picoline
Pronamide
206-44-0
86-73-7
321-60-8
367-12-4
76-44-8
1024-57-3
118-74-1
87-68-3
77-47-4
67-72-1
193-39-5
78-59-1
72-43-5
56-49-5
66-27-3
91-57-6
95-48-7
106-44-5
91-20-3
1146-65-2
134-32-7
91-59-8
88-74-4
99-09-2
100-01-6
98-95-3
4165-60-0
88-75-5
100-02-7
924-16-3
62-75-9
86-30-6
621-64-7
100-75-4
608-93-5
82-68-8
87-86-5
198-55-0
62-44-2
85-01-8

108-95-2
13127-88-3
109-06-8
23950-58-5
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
OS(44)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
DC(28)
DC(28)
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
NO
X
X
ND
ND
X
X
X
X
X
X
X
ND
X
X
X
ND
ND
ND
X
X
ND
X
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
X
X
X
X
ND
X
X
X
ND
ND
ND
X
X
ND
X
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
X
X
ND
ND
X
X
X
X
X
X
X
ND
X
X
X
ND
ND
ND
X
X
ND
X
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
                                  8250A  - 3
                                Revision 1
                            September 1994

-------
                                            Appropriate Preparation Techniques

Compounds                           CAS No"   3510     3520  3540/   3550  3580
                                                            3541
Pyrene
Terphenyl -d14(surr. )
1,2,4,5-Tetrachlorobenzene
2,3,4,6-Tetrachlorophenol
Toxaphene
2,4,6-Tribromopheno'l (surr. )
1,2,4-Trichlorobenzene
2 , 4 , 5-Tri chl orophenol
2,4,6-Trichlorophenol
129-00-0
1718-51-0
95-94-3
58-90-2
8001-35-2
118-79-6
120-82-1
95-95-4
88-06-2
X
X
X
X
X
X
X
X
X
X
X
ND
ND
X
X
X
X
X
X
ND
ND
ND
X
X
X
ND
X
X
X
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
8     Chemical Abstract Service Registry Number.

CP    =     Nonreproducible chromatographic performance.
DC    =     Unfavorable  distribution  coefficient  (number  in parenthesis  is
            percent recovery).
ND    =     Not determined.
OS    =     Oxidation  during   storage  (number  in   parenthesis  is  percent
            stability).
X     =     Greater than 70 percent  recovery by this technique.


      1.2   Method 8250  can be used to quantitate most  neutral,  acidic,  and
basic organic compounds that  are  soluble  in  methylene  chloride and capable of
being eluted without derivatization  as sharp  peaks from a gas chromatographic
packed  column.    Such  compounds  include  polynuclear   aromatic  hydrocarbons,
chlorinated  hydrocarbons   and  pesticides,  phthalate  esters,  organophosphate
esters,  nitrosamines,  haloethers,   aldehydes,   ethers,   ketones,  anilines,
pyridines,  quinolines,  aromatic  nitro  compounds,   and  phenols,  including
nitrophenols.  See Table 1  for a list of compounds  and their characteristic ions
that have been evaluated on the specified GC/MS system.

      1.3   The  following  compounds may require  special  treatment when being
determined by this method.   Benzidine can  be  subject to  oxidative losses during
solvent  concentration.   Also,  chromatography  is poor.   Under  the alkaline
conditions of the extraction step, a-BHC, 7-BHC, endosulfan I and  II,  and endrin
are subject to decomposition.  Neutral extraction  should be performed if these
compounds  are  expected   and   are   not  being  determined  by  Method  8080.
Hexachlorocyclopentadiene is subject  to thermal  decomposition in the inlet of the
gas chromatograph,  chemical  reaction  in  acetone  solution,  and  photochemical
decomposition.  N-nitrosodimethylamine is difficult to separate from the solvent
under  the  chromatographic   conditions  described.     N-nitrosodiphenylamine
decomposes  in the  gas chromatographic inlet  and cannot  be separated  from
diphenylamine. Pentachlorophenol,  2,4-dinitrophenol, 4-nitrophenol, 4,6-dinitro-
2-methylphenol,   4-chloro-3-methylphenol,   benzoic    acid,   2-nitroaniline,


                                   8250A -  4                          Revision  1
                                                                September 1994

-------
3-nitroaniline,  4-chloroaniline,  and  benzyl  alcohol  are  subject  to erratic
chromatographic behavior, especially if the GC system is contaminated with high
boiling material.

      1.4   The  estimated   quantitation   limit   (EQL)   of  Method  8250  for
determining an  individual  compound  is approximately  1 mg/kg  (wet  weight) for
soil/sediment samples, 1-200 mg/kg for wastes (dependent on  matrix and method of
preparation), and 10 /ug/L for ground water samples  (see Table  2).  EQLs will be
proportionately  higher for  sample  extracts  that  require  dilution  to  avoid
saturation of the detector.

      1.5   This method  is restricted to  use  by or  under  the  supervision of
analysts experienced  in  the  use  of gas chromatograph/mass  spectrometers and
skilled in  the  interpretation of mass spectra.  Each analyst  must demonstrate the
ability to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   Prior  to  using  this  method,  the samples should be prepared for
chromatography using  the  appropriate sample preparation and  cleanup  methods.
This  method describes  chromatographic  conditions   that  will  allow  for  the
separation of the compounds in the extract.


3.0   INTERFERENCES

      3.1   Raw  GC/MS data  from  all  blanks,   samples,  and  spikes  must  be
evaluated for interferences.  Determine if  the source of interference is in the
preparation and/or  cleanup of the samples and take corrective action to eliminate
the problem.

      3.2   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.   To reduce carryover, the
sample syringe must be rinsed out  between samples with solvent.   Whenever an
unusually  concentrated  sample is encountered,  it  should  be followed by the
analysis of solvent to check for cross contamination.


4.0   APPARATUS AND MATERIALS

      4.1   Gas chromatograph/mass spectrometer system

            4.1.1 Gas  chromatograph - An  analytical  system  complete with  a
      temperature-programmable  gas   chromatograph    suitable   for   splitless
      injection and all  required accessories,  including syringes,  analytical
      columns,  and  gases.

            4.1.2 Columns

                  4.1.2.1     For base/neutral  compound detection  -  2 m  x  2
            mm ID stainless or glass,  packed with 3%  SP-2250-DB on 100/120 mesh
            Supelcoport or equivalent.


                                  8250A -  5                         Revision  1
                                                                September 1994

-------
                  4.1.2.2      For acid compound detection  - 2 m x 2 mm ID glass,
            packed with 1% SP-1240-DA on 100/120 mesh Supelcoport or equivalent.

            4.1.3 Mass  spectrometer  -  Capable of scanning from 35  to  500 amu
      every 1 second or  less,  using  70  volts  (nominal)  electron energy in the
      electron impact ionization mode.  The mass  spectrometer must  be  capable
      of  producing  a mass  spectrum  for  decafluorotriphenylphosphine  (DFTPP)
      which meets all of the criteria in Table 3 when  1  /zL of the GC/MS tuning
      standard is injected through the GC (50 ng of DFTPP).

            4.1.4 GC/MS interface - Any GC-to-MS interface  that gives acceptable
      calibration points at 50 ng per  injection  for each  compound  of interest
      and achieves acceptable  tuning performance criteria may  be  used.  GC-to-MS
      interfaces  constructed  entirely of glass  or glass-lined materials are
      recommended.      Glass   may    be   deactivated   by    silanizing   with
      dichlorodimethylsilane.

            4.1.5 Data system - A computer system must  be interfaced  to the mass
      spectrometer.   The  system must  allow  the  continuous acquisition and
      storage on machine-readable media  of all mass spectra obtained throughout
      the  duration  of the  chromatographic  program.    The computer  must  have
      software that can search any GC/MS data file for ions  of a specific mass
      and that can plot  such  ion abundances versus time or scan number.   This
      type  of  plot  is  defined  as  an Extracted  Ion  Current  Profile  (EICP).
      Software must also be available that allows  integrating  the abundances in
      any EICP between specified time or  scan-number  limits.   The  most recent
      version of the EPA/NIH Mass Spectral Library should  also be available.

      4.2   Syringe  - 10 pi.


5.0   REAGENTS

      5.1   Reagent, grade chemicals shall  be used in all  tests. Unless otherwise
indicated, it is intended that all reagents shall conform to the specifications
of the Committee on Analytical  Reagents  of the American Chemical Society, where
such specifications are available. Other grades may be used, provided  it is  first
ascertained that the reagent  is  of sufficiently  high  purity  to  permit  its use
without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All references  to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Stock  standard solutions (1000  mg/L) - Standard  solutions  can be
prepared from pure standard materials or purchased as  certified solutions.

            5.3.1  Prepare  stock standard solutions by  accurately weighing  about
      0.0100 g  of pure  material.  Dissolve the material  in  pesticide  quality
      acetone  or other  suitable solvent  and  dilute to volume  in  a  10  ml
      volumetric  flask.   Larger  volumes  can be  used at  the  convenience of the
      analyst.  When compound  purity  is assayed to be 96% or greater, the weight
      may be used without correction to calculate the concentration of the  stock
      standard.   Commercially prepared stock  standards  may  be  used  at any


                                  8250A - 6                         Revision 1
                                                                September 1994

-------
      concentration  if  they  are  certified   by  the  manufacturer  or  by  an
      independent  source.

             5.3.2  Transfer the stock standard  solutions into bottles with Teflon
      lined  screw-caps or crimp  tops.   Store  at  -10°C to  -20°C  or less  and
      protect from light.  Stock standard solutions should be checked frequently
      for  signs  of degradation  or  evaporation,  especially  just  prior  to
      preparing calibration  standards from them.

             5.3.3  Stock  standard  solutions must  be replaced after 1 year  or
      sooner  if comparison  with  quality  control  check samples  indicates a
      problem.

      5.4    Internal standard solutions - The internal standards  recommended are
l,4-dichlorobenzene-d4,   naphthalene-d8,   acenaphthene-d10,   phenanthrene-d10,
cnrysene-d12, and perylene-d12.  Other compounds may be used as internal  standards
as long as  the requirements given in Sec. 7.3.2 are met.  Dissolve 200 mg of  each
compound with a  small volume of carbon disulfide. Transfer to a 50 ml volumetric
flask and dilute to volume with methylene chloride  so that the final  solvent is
approximately 20% carbon disulfide.  Most of the compounds  are  also  soluble in
small volumes of methanol, acetone, or toluene, except  for  perylene-d12.   The
resulting solution will contain each standard at a concentration  of 4,000 ng//nL.
Each 1 ml sample extract  undergoing  analysis should  be spiked with 10  juL  of the
internal standard  solution,  resulting in a concentration of 40 ng//iL of  each
internal standard.  Store  at -10°C  to -20°C or  less when not being  used.

      5.5   GC/MS  tuning standard  - A  methylene  chloride solution  containing
50 ng//LtL  of decafluorotriphenylphosphine  (DFTPP)  should  be  prepared.    The
standard should also contain 50 ng/juL each of 4,4'-DDT,  pentachlorophenol,  and
benzidine to verify injection port  inertness and GC column  performance.  Store
at 4°C or less  when not being used.

      5.6   Calibration  standards - Calibration standards at a  minimum of  five
concentrations should  be prepared.  One of the  calibration  standards  should  be
at a concentration  near, but above, the method detection limit; the others  should
correspond to the range of concentrations found in real  samples but  should  not
exceed the working  range  of the GC/MS system.   Each  standard  should contain  each
analyte for detection  by this method (e.g. some or all of the compounds  listed
in Table 1 may be included).  Each  1 ml aliquot of calibration  standard  should
be spiked with 10 juL of the  internal standard solution prior to analysis.   All
standards should be stored at -10°C  to -20°C and should be freshly prepared  once
a year,  or sooner if check standards indicate  a  problem.  The daily  calibration
standard should be prepared weekly  and stored at 4°C.

      5.7   Surrogate  standards  -   The  recommended  surrogate standards  are
phenol-de,    2-fluorophenol,    2,4,6-tribromophenol,    nitrobenzene-d5,     2-
fluorobiphenyl,  and p-terphenyl-d14.  See Method 3500  for  the  instructions  on
preparing the surrogate standards.  Determine what concentration should be in the
blank extracts after all  extraction, cleanup,  and concentration  steps.   Inject
this concentration into the GC/MS  to determine recovery of surrogate  standards
in all blanks,  spikes,  and sample  extracts.  Take into account all dilutions  of
sample extracts.
                                   8250A  -  7                         Revision 1
                                                                September 1994

-------
      5.8   Matrix  spike  standards  -  See Method  3500  for  instructions  on
preparing the matrix spike standard.  Determine what concentration should be in
the  blank  extracts  after  all  extraction,  cleanup,  and  concentration  steps.
Inject this concentration into the GC/MS to determine recovery of standards in
all matrix spikes.  Take into account all dilutions of sample extracts.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See  the  introductory material to this chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Sample  preparation  -  Samples  must  be  prepared  by  one  of  the
following methods prior to GC/MS analysis.

      Matrix                                Methods
      Water                                 3510, 3520
      Soil/sediment                         3540, 3541, 3550
      Waste                                 3540, 3541, 3550, 3580

            7.1.1 Direct  injection  -  In  very  limited  applications  direct
      injection of the sample into  the GC/MS system with a 10 p.1 syringe may be
      appropriate.     The  detection   limit   is  very   high   (approximately
      10,000 Mg/L);  therefore,  it  is  only  permitted where  concentrations  in
      excess of  10,000 /Ltg/L are expected.  The system must be  calibrated  by
      direct injection.

      7.2   Extract cleanup - Extracts may be cleaned up by any of the following
methods prior to GC/MS analysis.

      Compounds                                 Methods
      Phenols                                   3630,  3640,  8040a
      Phthalate esters                          3610,  3620,  3640
      Nitrosamines                              3610,  3620,  3640
      Organochlorine pesticides & PCBs          3620,  3640,  3660
      Nitroaromatics and cyclic ketones         3620,  3640
      Polynuclear aromatic hydrocarbons         3611,  3630,  3640
      Haloethers                                3620,  3640
      Chlorinated hydrocarbons                  3620,  3640
      Organophosphorus pesticides               3620
      Petroleum waste                           3611,  3650
      All basic, neutral, and acidic
      Priority Pollutants                       3640


aMethod  8040 includes a derivatization technique followed by  GC/ECD analysis, if
interferences are encountered on GC/FID.
                                   8250A -  8                         Revision 1
                                                                September 1994

-------
7.3   Recommended GC/MS  operating conditions

Electron energy:               70 volts (nominal)
Mass range:                    35-500 amu
Scan time:                     1  sec/scan
Injector temperature:          250-300°C
Transfer line temperature:     250-300°C
Source temperature:            According  to manufacturer's specifications
Injector:                      Grob-type, splitless
Sample volume:                 1-2 Mi-
Carrier gas:                   Helium at  30 mL/min

Conditions for base/neutral analysis  (3% SP-2250-DB):

Initial column temperature and hold time:       50°C for 4 minutes
Column temperature program:                     50-300°C at 8°C/min
Final column temperature hold:                  300°C for 20 minutes

Conditions for acid analysis  (1% SP-1240-DA):

Initial column temperature and hold time:       70°C for 2 minutes
Column temperature program:                     70-200°C at 8°C/min
Final column temperature hold:                  200°C for 20 minutes

7.4 Initial calibration

      7.4.1 Each GC/MS system must be hardware-tuned to  meet the criteria
in Table 3  for  a 50 ng  injection of DFTPP.   Analyses  should  not begin
until  all  these  criteria  are met.    Background  subtraction  should  be
straightforward and designed only to  eliminate column bleed or instrument
background ions.  The GC/MS tuning standard should also  be used to assess
GC column performance and  injection  port inertness.  Degradation of DDT
to DDE  and  ODD  should not exceed 20% (See  Sec.  7.4.5  of Method 8080).
Benzidine  and  pentachlorophenol should be  present   at  their  normal
responses,  and  no peak  tailing  should  be visible.  If  degradation  is
excessive and/or  poor  chromatography  is  noted,  the injection  port may
require cleaning.

      7.4.2 The internal  standards selected in Sec. 5.1 should permit most
of the components of interest in a chromatogram to have retention times
of 0.80-1.20 relative to  one of the internal  standards.   Use the base peak
ion  from  the  specific  internal  standard  as  the  primary  ion  for
quantitation (see Table 1).  If interferences are noted,  use the next most
intense ion as the quantitation  ion  (i.e. for l,4-dichlorobenzene-d4 use
m/z 152 for quantitation).
                            8250A  - 9                         Revision 1
                                                          September 1994

-------
      7.4.3 Analyze 1 /zL of each calibration standard (containing internal
standards) and tabulate the area of the primary characteristic  ion against
concentration  for  each compound  (as  indicated in  Table  1).   Calculate
response factors (RFs) for  each compound relative to the internal standard
as follows:

      RF - (AxCis)/(AisCJ

where:

Ax    =     Area  of  the  characteristic  ion  for  the  compound  being
            measured.
Ais    =     Area  of  the  characteristic ion  for the  specific internal
            standard.
Cx    =     Concentration  of the  compound  being measured  (ng//iL).
Cjs    =     Concentration  of the  specific  internal  standard  (ng/juL).

      7.4.4 A  system  performance  check must be performed to ensure that
minimum  average  response  factors,  calculated  as  the  mean  of  the  5
individual relative response factors, are met before the calibration curve
is  used.    For semivolatiles,  the System Performance  Check Compounds
(SPCCs)  are:    N-nitroso-di-n-propylamine;  hexachlorocyclopentadiene;
2,4-dinitrophenol; and 4-nitrophenol.  The minimum  acceptable  average RF
for these  compounds  is 0.050.   These  SPCCs  typically  have very low RFs
(0.1-0.2) and tend to decrease in response as the chromatographic system
begins to  deteriorate or  the  standard material  begins  to deteriorate.
They are usually the  first  to show poor performance.  Therefore, they must
meet the minimum requirement when the  system  is calibrated.

            7.4.4.1      The percent relative standard deviation should be
      less  than 15%  for  each  compound.   However, the %RSD  for each
      individual Calibration Check  Compound (CCC) (see Table 4) must be
      less than  30%.   The relative retention  times of each compound in
      each calibration run should agree within  0.06 relative retention
      time  units.    Late-eluting compounds  usually  have much  better
      agreement.
                         SD
            %RSD = 	—	   x 100
                         RF
      where:
             RSD    =      relative  standard  deviation.
             RF     =      mean  of 5 initial  RFs  for  a  compound.
             SD     =      standard  deviation of average RFs for a compound.
                            8250A - 10                        Revision  1
                                                          September  1994

-------
            SD =
 N  (RFj - RF)
 I  	
1=1 N - 1
                                        2
      where:
            RFj    =      RF  for  each  of  the  5  calibration  levels
            N      =      Number  of  RF values  (i.e.,  5)

            7.4.4.2      If  the %RSD of any CCC is 30% or greater, then the
      chromatographic system is  too reactive for analysis to begin.  Clean
      or replace the injector liner  and/or capillary column, then repeat
      the calibration procedure beginning with Sec. 7.4.

      7.4.5 Linearity -  If  the %RSD  of any compound is  15% or less, then
the  relative  response  factor   is  assumed   to  be  constant  over  the
calibration range, and the  average relative  response  factor may be used
for quantitation (Sec. 7.7.2).

            7.4.5.1      If  the %RSD  of any compound is greater than 15%,
      construct   calibration   curves   of   area  ratio   (A/Ais)   versus
      concentration using first  or  higher order regression fit of the five
      calibration points.  The analyst should  select the regression order
      which introduces the  least calibration error into the quantitation
      (Sees. 7.7.2.2  and 7.7.2.3).  The use  of  calibration curves is a
      recommended alternative to average response factor calibration, and
      a useful  diagnostic of standard preparation accuracy and absorption
      activity in the chromatographic system.

7.5   Daily GC/MS calibration

      7.5.1 Prior to analysis of samples,  the GC/MS tuning standard must
be analyzed.  A 50 ng  injection  of DFTPP must result  in a mass spectrum
for DFTPP which meets the criteria  given in Table 3.  These criteria must
be demonstrated during each 12 hour  shift.

      7.5.2 A calibration standard(s) at mid-concentration containing all
semivolatile  analytes,   including  all   required  surrogates,   must  be
analyzed every  12  hours during analysis.  Compare the instrument response
factor from the standards every  12 hours with  the  SPCC (Sec. 7.5.3) and
CCC (Sec. 7.5.4) criteria.

      7.5.3 System  Performance  Check  Compounds  (SPCCs)  -  A  system
performance check must be made during every  12  hour shift.   If the SPCC
criteria  are  met, a  comparison  of  response  factors   is  made for  all
compounds.  This  is  the  same check  that  is applied during  the  initial
calibration. If the minimum response factors are not met, the system must
be evaluated,  and  corrective action must be taken before sample analysis
begins.   The minimum RF  for semivolatile SPCCs  is  0.050.   Some possible
problems  are   standard   mixture  degradation,  injection   port   inlet
contamination,  contamination  at the  front end  of the  analytical  column,
                            8250A - 11                        Revision 1
                                                          September 1994

-------
and active sites in the column or chromatographic system.  This check must
be met before analysis begins.

      7.5.4 Calibration  Check  Compounds  (CCCs):    After  the  system
performance check is  met,  CCCs  listed in Table 4 are  used  to  check the
validity of the initial calibration.

      Calculate the percent drift using:

                C   -  C
      % Drift = - - -  x 100
where:

      C,  =   Calibration Check Compound standard concentration.
      Cc =  Measured concentration using selected quantitation method.

      If the percent difference  for each CCC is less  than  or equal to 20%,
the initial calibration is assumed to be valid.  If the criterion is not
met  (>  20%  drift)  for  any one  CCC,  corrective  action must  be taken.
Problems similar to those  listed under SPCCs could affect this criterion.
If no source of the problem can  be determined after corrective action has
been  taken,  a  new  five-point   calibration  must  be  generated.    This
criterion must be met before sample  analysis begins.  If the CCCs are not
analytes required by the permit, then all required analytes must meet the
20% drift criterion.

      7.5.5 The  internal  standard responses  and retention times  in the
calibration check standard must be evaluated  immediately after or during
data acquisition.  If the retention time for any internal standard changes
by more than 30  seconds from the last  daily calibration  (Sec.  7.4), the
chromatographic system must be  inspected for malfunctions and corrections
must  be  made.,  as required.   If the EICP  area for  any  of the internal
standards changes by a factor of two (-50% to  +100%) from the last daily
calibration check standard, the mass spectrometer must be  inspected for
malfunctions and corrections must be made,  as  appropriate.

7.6   GC/MS analysis

      7.6.1 It  is  highly  recommended that  the  extract be  screened  on a
GC/FID or  GC/PID using the same type  of  column.   This will  minimize
contamination of the GC/MS  system from  unexpectedly high concentrations
of organic compounds.

      7.6.2 Spike the  1 ml extract obtained from sample preparation with
10 juL of the internal standard solution (Sec. 5.4) just  prior to analysis.

      7.6.3 Analyze the 1  ml extract by GC/MS using the  appropriate column
(as specified in Sec.  4.1.2).  The recommended GC/MS operating conditions
to be used are specified in Sec.  7.3.
                            8250A - 12                        Revision 1
                                                          September 1994

-------
      7.6.4 If the response for any quantitation ion exceeds the initial
calibration curve range of the GC/MS  system,  extract  dilution must take
place.  Additional  internal  standard must be added to the diluted extract
to maintain  the required  40  ng/juL  of  each  internal  standard  in  the
extracted volume.  The diluted extract must be reanalyzed.

      7.6.5 Perform  all  qualitative  and  quantitative  measurements  as
described in Sec. 7.7.  Store the extracts at 4°C,  protected from  light
in screw-cap vials  equipped with unpierced Teflon lined septa.

7.7   Data interpretation

      7.7.1 Qualitative analysis

            7.7.1.1     The   qualitative   identification   of  compounds
      determined by   this  method  is  based on  retention  time,  and  on
      comparison of the sample mass spectrum, after background correction,
      with  characteristic   ions   in  a  reference mass  spectrum.    The
      reference mass  spectrum must be generated  by  the  laboratory using
      the conditions  of this  method.   The characteristic  ions  from the
      reference mass  spectrum  are defined to be the three ions of greatest
      relative intensity,  or any  ions over 30% relative intensity if less
      than three such ions  occur in  the reference  spectrum.   Compounds
      should be identified as present when the criteria below are met.

                  7.7.1.1.1   The  intensities of the characteristic ions
            of a compound maximize in the same scan or within one scan of
            each other.   Selection of  a  peak by  a data  system  target
            compound  search  routine  where  the  search  is  based  on  the
            presence  of  a  target  chromatographic  peak containing  ions
            specific  for  the  target  compound   at  a  compound-specific
            retention time will be accepted as meeting this criterion.

                  7.7.1.1.2   The  RRT of the  sample component is  within
            ± 0.06 RRT units of the RRT of the standard component.

                  7.7.1.1.3   The    relative    intensities     of    the
            characteristic  ions   agree  within  30%  of   the  relative
            intensities  of  these  ions   in  the   reference  spectrum.
            (Example:   For  an  ion  with  an  abundance  of  50%  in  the
            reference spectrum, the corresponding abundance  in  a  sample
            spectrum  can range between 20% and 80%.)

                  7.7.1.1.4   Structural   isomers   that   produce   very
            similar  mass  spectra  should  be  identified  as  individual
            isomers  if  they  have sufficiently  different GC  retention
            times.   Sufficient GC  resolution is achieved if the height of
            the valley between two isomer peaks  is  less than 25%  of the
            sum of the two peak heights.   Otherwise,  structural isomers
            are identified  as isomeric pairs.

                  7.7.1.1.5   Identification  is  hampered  when  sample
            components are not resolved chromatographically  and produce


                            8250A  - 13                         Revision 1
                                                          September 1994

-------
      mass  spectra  containing ions contributed  by more  than  one
      analyte.  When gas chromatographic peaks obviously represent
      more than one sample  component  (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of   analyte  spectra  and  background
      spectra is important.   Examination  of  extracted  ion current
      profiles  of  appropriate ions  can aid  in  the selection  of
      spectra, and  in qualitative identification of compounds.  When
      analytes  coelute  (i.e.,  only  one  chromatographic peak  is
      apparent), the identification criteria  can  be  met,  but each
      analyte spectrum will contain extraneous ions contributed by
      the coeluting compound.

      7.7.1.2     For samples containing components not associated
with the calibration standards, a library search may be made for the
purpose of tentative  identification.  The necessity to perform this
type of  identification  will be determined by the purpose  of  the
analyses  being  conducted.    Computer  generated  library  search
routines  should   not   use  normalization  routines   that   would
misrepresent the library or  unknown spectra  when  compared to each
other.   For  example, the RCRA permit or waste delisting requirements
may require the reporting of nontarget analytes.  Only after visual
comparison of sample  spectra with  the nearest library searches will
the  mass spectral   interpretation specialist  assign  a  tentative
identification. Guidelines  for making tentative identification are:

      (1)   Relative intensities  of  major ions  in  the  reference
spectrum (ions > 10%  of  the  most abundant ion) should be present in
the sample spectrum.

      (2)  The relative intensities of the major ions should agree
within + 20%.   (Example:  For an ion with an abundance of 50% in the
standard spectrum,  the  corresponding sample  ion  abundance must be
between 30 and 70%.)

      (3)  Molecular ions present in  the reference spectrum should
be present in sample the spectrum.

      (4)   Ions present  in the  sample spectrum  but  not  in  the
reference  spectrum should  be  reviewed  for  possible  background
contamination or presence of coeluting compounds.

      (5)   Ions  present in the reference spectrum but  not  in  the
sample spectrum should be reviewed for possible subtraction from the
sample  spectrum  because of background  contamination or coeluting
peaks.   Data system library  reduction programs can sometimes create
these discrepancies.

7.7.2 Quantitative Analysis

      7.7.2.1     When   a   compound   has  been   identified,   the
quantitation  of that  compound  will   be based  on the  integrated
abundance from the EICP of the primary characteristic ion.
                      8250A - 14                        Revision 1
                                                    September 1994

-------
      7.7.2.2      If the %RSD  of a compound's  relative response
factor is 15% or less, then the  concentration in  the  extract  may be
determined  using the  average  response  factor  (RF)  from initial
calibration data  (Sec. 7.4.3) and the following  equation:

                     (Ax x Cis)
      Cex (mg/L)  =
                     (Ais x RF)

      where Cex  is the concentration of the compound in the extract,
and the other terms  are as defined in Sec. 7.4.3.

      7.7.2.3     Alternatively,  the regression  line fitted to the
initial calibration  (Sec.  7.4.6.1) may  be used  for  determination of
the extract concentration.

      7.7.2.4     Compute the  concentration  of the analyte in the
sample using the equations in Sees. 7.7.2.4.1  and  7.7.2.4.2.

            7.7.2.4.1   The  concentration of  the  analyte  in the
      liquid  phase   of  the  sample   is  calculated   using  the
      concentration  of the analyte in the extract and the volume of
      liquid extracted, as follows:

            Concentration in liquid (^g/L) = |Cex x VeJ
                                                  V.
                                                   o
      where:

            Vex    =      extract  volume,  in  ml
            V0     =      volume of  liquid extracted,  in L.

            7.7.2.4.2    The  concentration  of  the analyte  in the
      solid   phase  of  the   sample   is   calculated   using  the
      concentration of the pollutant in  the extract and the weight
      of the solids, as  follows:

            Concentration in solid  (/ug/kg)  =  (Cex x VeK)


      where:

            Vex    =      extract  volume,  in  mL
            Ws     =      sample weight,  in kg.

      7.7.2.5      Where applicable, an estimate of concentration for
noncalibrated components in the sample should be made.   The formulae
given above should be used with  the following  modifications:  The
areas Ax and Ais should  be from  the  total  ion chromatograms and the
RF for the  compound should be  assumed  to be 1.  The concentration
obtained should  be reported  indicating  (1)  that the  value  is  an
estimate and  (2)  which  internal standard  was used  to  determine
                      8250A - 15                        Revision 1
                                                    September 1994

-------
            concentration.    Use   the   nearest   internal   standard  free  of
            interferences.

                  7.7.2.6     Quantitation  of multicomponent  compounds  (e.g.
            Aroclors)  is  beyond  the  scope  of  Method  8250A.    Normally,
            quantitation is performed using a GC/ECD by Method 8080.


8.0   QUALITY CONTROL

      8.1   Each  laboratory  that  uses these methods is required  to operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory capability and an ongoing analysis of
spiked samples  to evaluate  and document data quality.   The  laboratory  must
maintain records  to document  the quality of the  data  generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet  the performance characteristics of the method.
When results of sample spikes indicate  atypical  method performance, a quality
control check standard must be  analyzed  to  confirm  that the measurements were
performed in an in-control mode  of operation.

      8.2   Before  processing  any  samples,  the  analyst should  demonstrate,
through the  analysis  of  a reagent  water blank,  that  interferences  from the
analytical system, glassware, and reagents are under control.  Each time a set
of samples is extracted  or there is  a  change  in reagents, a reagent water blank
should be  processed as a  safeguard against chronic laboratory  contamination.  The
blank samples should be carried  through all  stages of the sample preparation and
measurement steps.

      8.3   The   experience  of the  analyst  performing  GC/MS  analyses  is
invaluable to the success  of the methods.  Each day that analysis is performed,
the  daily  calibration  standard  should  be  evaluated  to   determine  if  the
chromatographic system is  operating properly.  Questions  that should be asked
are:   Do  the peaks look  normal?;  Is  the response obtained  comparable  to the
response  from previous  calibrations?   Careful   examination  of  the  standard
chromatogram  can  indicate whether the column  is still good,  the  injector is
leaking, the injector septum needs replacing, etc.  If any changes are made to
the system (e.g.  column changed),  recalibration of the system must take place.

      8.4   Required  instrument QC  is found  in the following  section:

            8.4.1 The   GC/MS   system  must   be   tuned  to   meet   the   DFTPP
      specifications  in Sec. 7.3.1 and 7.4.1.

            8.4.2 There must  be an  initial  calibration of the GC/MS system as
      specified in Sec.  7.4.

            8.4.3 The GC/MS system must meet the SPCC criteria specified in Sec.
      7.5.3 and the CCC criteria in Sec. 7.5.4, each 12 hr.

      8.5   To  establish  the  ability  to  generate  acceptable   accuracy  and
precision, the analyst must perform the following operations.
                                  8250A - 16                        Revision 1
                                                                September 1994

-------
             8.5.1  A quality control  (QC) check sample concentrate is required
      containing each  analyte at a concentration of 100 mg/L in acetone.  The
      QC check  sample  concentrate  may be prepared from pure standard materials
      or purchased  as  certified  solutions.   If prepared by the laboratory, the
      QC check  sample  concentrate must  be  made using stock standards prepared
      independently from  those used for  calibration.

             8.5.2  Using a pipet, prepare QC check samples  at  a concentration of
      100 jug/L  by adding  1.00 ml of QC check sample  concentrate to  each  of four
      1-L aliquots  of  organic-free reagent water.

             8.5.3  Analyze the well-mixed  QC check  samples  according  to the
      method beginning in Sec. 7.1 with  extraction of the samples.

             8.5.4  Calculate the  average  recovery  (x) in /zg/L, and  the standard
      deviation of  the recovery  (s)  in /xg/L, for each  analyte  using the four
      results.

             8.5.5  For  each  analyte  compare  s  and  x  with  the  corresponding
      acceptance criteria jfor precision and  accuracy,  respectively,  found in
      Table  6.   If  s  and x for  all  analytes of  interest  meet  the acceptance
      criteria, the system  performance  is  acceptable and analysis  of actual
      samples can begin.   If any  individual s exceeds the precision  limit or any
      individual  x  falls outside the  range  for accuracy,  then the  system
      performance  is unacceptable  for that analyte.

             NOTE:  The  large  number of analytes in Table  6 present a substantial
                   probability that one  or  more will fail  at  least one of the
                   acceptance criteria when  all  analytes of a given method are
                   analyzed.

             8.5.6 When one  or more of the analytes tested fail at  least one of
      the  acceptance criteria,  the  analyst  must  proceed according  to Sees.
      8.5.6.1 or 8.5.6.2.

                  8.5.6.1     Locate  and correct  the source of the problem and
             repeat  the  test for  all  analytes of  interest  beginning  with Sec.
             8.5.2.

                  8.5.6.2     Beginning  with Sec. 8.5.2,  repeat  the  test only
             for those analytes that failed  to meet criteria.  Repeated failure,
             however, will  confirm a general problem with  the measurement  system.
             If  this occurs,  locate and correct the  source of  the  problem and
             repeat  the test  for  all  compounds of interest beginning  with Sec.
             8.5.2.

      8.6    The laboratory  must,  on  an  ongoing  basis,  analyze  a  method blank,
a matrix spike,  and a matrix spike/duplicate  for  each analytical batch (up to a
maximum of 20 samples/batch)  to assess accuracy.   For  laboratories analyzing one
to ten samples per month,  at least one spiked sample per month is required.
                                  8250A - 17                        Revision 1
                                                                September 1994

-------
      8.6.1 The  concentration  of, the  spike  in  the  sample  should  be
determined as follows:

            8.6.1.1      If, as in compliance monitoring, the concentration
      of  a  specific analyte  in the sample  is being checked  against  a
      regulatory concentration  limit, the  spike  should  be at  that limit
      or 1 to 5 times higher than the background concentration determined
      in Sec. 8.6.2, whichever  concentration would be larger.

            8.6.1.2      If the concentration of a specific analyte in the
      sample  is not  being checked  against  a limit  specific to  that
      analyte, the spike should be at 100 /^g/L or 1 to 5 times higher than
      the background  concentration  determined in  Sec.  8.6.2,  whichever
      concentration would be larger.

            8.6.1.3      If  it  is  impractical  to determine  background
      levels  before  spiking   (e.g.,  maximum  holding  times  will  be
      exceeded), the spike concentration should be at (1) the  regulatory
      concentration limit, if  any; or,  if  none (2) the larger of either
      5  times  higher  than  the  expected   background concentration  or
      100 /xg/L.

      8.6.2 Analyze  one  sample  aliquot   to  determine  the  background
concentration (B) of each analyte.  If necessary, prepare a new QC check
sample   concentrate   (Sec.   8.5.1)   appropriate  for   the   background
concentration in the sample.   Spike a second sample aliquot with 1.00 ml
of the QC  reference  sample concentrate and  analyze  it  to determine the
concentration after spiking (A) of each analyte.  Calculate each percent
recovery (p) as 100(A-B)%/T, where T is the  known true value of the spike.

      8.6.3 Compare  the percent recovery  (p)  for  each  analyte  with the
corresponding QC acceptance criteria found  in Table 6.  These  acceptance
criteria were calculated to include  an allowance  for  error in measurement
of both  the  background  and  spike concentrations,  assuming  a  spike  to
background ratio of 5:1.  This error will  be accounted for to  the extent
that the analyst's spike to background ratio approaches 5:1.   If spiking
was performed at a concentration lower than  100 /ug/U  the  analyst must use
either the  QC acceptance criteria presented in Table 6,  or optional  QC
acceptance criteria calculated for the specific spike concentration.  To
calculate optional  acceptance  criteria for the recovery of an analyte: (1)
Calculate accuracy (x')  using  the  equation  found  in Table 7, substituting
the spike concentration  (T) for C;  (2)  calculate overall precision (S')
using the equation in Table 7,  substituting x' for x; (3) calculate the
range   for   recovery   at   the   spike   concentration   as   (100x'/T)
+ 2.44(100S'/T)%.

      8.6.4  If  any  individual  p falls outside  the designated range for
recovery,  that  analyte  has  failed  the  acceptance  criteria.   A check
standard containing each analyte that failed the criteria  must be analyzed
as described  in Sec. 8.7.
                            8250A - 18                        Revision 1
                                                          September 1994

-------
      8.7    If  any  analyte  fails  the acceptance criteria for recovery in Sec.
8.6, a QC check  standard containing each analyte  that failed must be  prepared and
analyzed.

      NOTE:  The frequency for the required analysis of a QC check standard will
             depend upon the number of analytes being simultaneously tested, the
             complexity  of  the  sample matrix,  and  the  performance of  the
             laboratory.   If the  entire list of  analytes  in Table  6 must be
             measured  in the  sample  in Sec.  8.6, the probability  that  the
             analysis of a QC check standard will be required is high.  In this
             case, the QC check standard should  be routinely analyzed with the
             spiked sample.

             8.7.1 Prepare the  QC  reference sample by  adding 1.0 mL  of the QC
      check  sample concentrate  (Sec.  8.5.1  or 8.6.2)  to 1  L of reagent water.
      The QC check standard  needs  only to  contain  the analytes  that failed
      criteria  in the test  in Sec. 8.6.

             8.7.2 Analyze the  QC  check standard to determine the concentration
      measured  (A)  of  each  analyte.    Calculate  each  percent recovery (Ps) as
      100(A/T)%, where T is the true value of the  standard concentration.

             8.7.3 Compare the  percent recovery  (PJ for each analyte with the
      corresponding QC acceptance  criteria  found in Table 6.   Only analytes that
      failed the test in Sec. 8.6 need  to be compared with these criteria.  If
      the recovery of any such analyte  falls outside the designated range, the
      laboratory performance for  that  analyte  is  judged to  be  out  of control,
      and the problem must  be immediately  identified and corrected.  The result
      for that analyte in the unspiked  sample is suspect and may  not be reported
      for regulatory compliance purposes.

      8.8    As  part of  the  QC  program for the  laboratory,  method accuracy for
each matrix studied must be  assessed  and records must  be maintained.  After the
analysis of  five spiked samplesJof the same matrix) as  in Sec. 8.6,  calculate
the average  percent recovery  (p)  and  the  standard  deviation   of  the percent
recovery (sp).   Express  the  accuracy  assessment  as a percent recovery interval
from p  - 2sp to p +  2sp.   If p = 90%  and  sp =  10%, for example,  the accuracy
interval is  expressed  as 70-110%.   Update  the accuracy assessment  for each
analyte  on   a   regular  basis  (e.g.    after  each  five  to   ten  new  accuracy
measurements).

      8.9    To determine acceptable accuracy and precision limits for surrogate
standards  the following procedure should be performed.

             8.9.1 For each  sample analyzed,  calculate the  percent  recovery of
      each surrogate in the sample.

             8.9.2 Once a minimum of thirty samples of  the same matrix have been
      analyzed,   calculate   the  average  percent   recovery   (P)  and  standard
      deviation of the percent recovery (s) for each of the surrogates.
                                  8250A - 19                        Revision 1
                                                                September 1994

-------
            8.9.3  For  a given matrix,  calculate  the upper and  lower control
      limit for method performance for each  surrogate standard.  This should be
      done as follows:

            Upper  Control Limit (UCL) = P + 3s
            Lower  Control Limit (LCL) = P - 3s

            8.9.4  For  aqueous  and  soil  matrices,  these  laboratory established
      surrogate  control  limits should,  if applicable,  be compared  with  the
      control limits listed in Table 8.  The limits given in Table 8 are multi-
      laboratory performance  based limits for  soil  and aqueous  samples,  and
      therefore, the  single-laboratory  limits established in Step  8.9.3  must
      fall within  those given in Table 8 for these matrices.

            8.9.5  If recovery is not within limits, the following procedures are
      required.

            •      Check  to  be sure  there  are  no   errors  in  calculations,
                   surrogate  solutions  and  internal   standards.   Also,  check
                   instrument performance.

            •      Recalculate the  data  and/or reanalyze  the  extract if any of
                   the  above checks  reveal a problem.

            •      Reextract and reanalyze the sample if  none  of the above are
                   a problem or flag the data  as "estimated concentration".

            8.9.6  At a minimum, each laboratory should update surrogate recovery
      limits on a matrix-by-matrix basis,  annually.

      8.10  It  is  recommended that the  laboratory  adopt additional  quality
assurance practices for use with this method.   The specific practices that are
most productive depend upon the needs of  the  laboratory  and  the nature of the
samples.   Field duplicates may  be  analyzed  to  assess  the  precision  of  the
environmental measurements.  When doubt exists over the identification  of a peak
on the chromatogram, confirmatory techniques such as gas chromatography with a
dissimilar column or mass spectrometry using other  ionization modes must be used.
Whenever possible,  the laboratory should  analyze standard reference materials and
participate in relevant performance evaluation studies.


9.0   METHOD PERFORMANCE

      9.1   Method 8250 was  tested  by 15 laboratories  using  organic-free reagent
water, drinking water, surface water, and industrial  wastewaters spiked at six
concentrations  over  the  range  5-1,300 M9/L.   Single  operator  accuracy  and
precision,  and method  accuracy  were   found  to  be  directly  related  to  the
concentration of the analyte and essentially independent of the sample matrix.
Linear equations to describe these relationships are presented in Table 7.
                                  8250A - 20                        Revision 1
                                                                September 1994

-------
10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act,  Method 625," October 26,
      1984.

2.    U.S.  EPA Contract  Laboratory Program,  Statement of  Work  for Organic
      Analysis, July  1985, Revision.

3.    Eichelberger, J.W., I.E. Harris,  and  W.L.  Budde, "Reference Compound to
      Calibrate   Ion  Abundance   Measurement   in  Gas   Chromatography-Mass
      Spectrometry Systems," Analytical Chemistry, 47, 995-1000, 1975.

4.    "Method Detection Limit for Methods  624  and  625," Olynyk,  P., W.L. Budde,
      and J.W. Eichelberger, Unpublished report, October 1980.

5.    "Inter!aboratory Method Study  for EPA Method 625-Base/Neutrals, Acids, and
      Pesticides," Final Report for EPA Contract  68-03-3102  (in preparation).

6.    Burke,  J.A.  "Gas  Chromatography  for Pesticide  Residue  Analysis;  Some
      Practical Aspects,"  Journal  of the  Association of  Official  Analytical
      Chemists, 48, 1037, 1965.
                                  8250A - 21                        Revision 1
                                                                September 1994

-------
                        TABLE 1.
CHROMATOGRAPHIC CONDITIONS, METHOD DETECTION LIMITS, AND
     CHARACTERISTIC IONS FOR SEMIVOLATILE COMPOUNDS
Compound
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
Aldrin
4-Aminobiphenyl
Aniline
Anthracene
Aroclor-1016b
Aroclor-1221b
Aroclor-1232b
Aroclor-1242b
Aroclor-1248b
Aroclor-1254b
Aroclor-1260b
Benzidine8
Benzoic acid
Benzo(a)anthracene
Benzo(b)fl uoranthene
Benzo(k)fluoranthene
Benzo (g , h , i ) peryl ene
Benzo(a)pyrene
Benzyl alcohol
a-BHCa
0-BHC

-------
TABLE 1.
(Continued)
Compound
4,4'-DDT
4,4'-DDE
Dibenz(a,j)acridine
Dibenz (a, h) anthracene
Dibenzofuran
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S.)
3 , 3 ' -Di chl orobenz i d i ne
2,4-Dichlorophenol
2,6-Dichlorophenol
Dieldrin
Diethyl phthalate
p-Dimethylaminoazobenzene
7, 12-Dimethylbenz (a) anthracene
a- , a-Di methyl phenethyl ami ne
2,4-Dimethylphenol
Dimethyl phthalate
4,6-Dinitro-2-methylphenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diphenylamine
1,2-Diphenylhydrazine
Di-n-octyl phthalate
Endosulfan I"
Endosulfan II"
Endosulfan sulfate
Endrin8
Endrin aldehyde
Endrin ketone
Ethyl methanesulfonate
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachl orobenzene
Hexachlorobutadiene
Hexachl orocycl opentadi enea
Hexachloroethane
Method
Retention Detection Primary Secondary
Time (min) Limit (p.g/1) Ion Ion(s)
29.3
27.2
--
43.2
--
24.7
8.4
7.4
7.8
--
32.2
9.8
--
27.2
20.1
--
--
--
9.4
18.3
16.2
15.9
19.8
18.7
--
--
32.5
26.4
28.6
29.8
27.9
--
--
--
26.5
19.5
--
--
23.4
25.6
21.0
11.4
13.9
8.4
4.7
--
--
2.5
--
2.5
1.9
1.9
4.4
--
16.5
2.7
--
2.5
1.9
--
--
--
2.7
1.6
24
42
5.7
1.9
--
--
2.5
--
--
5.6
--

--
--
2.2
1.9
--
--
1.9
2.2
1.9
0.9
--
1.6
235
246
279
278
168
149
146
146
146
152
252
162
162
79
149
120
256
58
122
163
198
184
165
165
169
77
149
195
337
272
263
67
317
79
202
166
172
112
100
353
284
225
237
117
237,
24,
280,
139,
139
150,
148,
148,
148,
150,
254,
164,
164,
263,
177,
225,
241,
91,
107,
194,
51,
63,
63,
63,
168,
105,
167,
339,
339,
387,
82,
345,
67,
109,
101,
165,
171
64
272,
355,
142,
223,
235,
201,
165
176
277
279

104
111
111
111
115
126
98
98
279
150
77
257
42
121
164
105
154
89
89
167
182
43
341
341
422
81
250
319
97
203
167


274
351
249
227
272
199
8250A - 23
    Revision 1
September 1994

-------
                                   TABLE  1.
                                  (Continued)
Compound
            Method
Retention   Detection
Time (min)  Limit (jug/L)
Primary  Secondary
Ion        Ion(s)
Indeno(l,2,3-cd)pyrene
Isophorone
Methoxychlor
3-Methyl chol anthrene
Methyl methanesulfonate
2-Methyl naphthalene
2-Methyl phenol
4-Methyl phenol
Naphthalene
Naphthalene-d8 (I.S.)
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
Nitrobenzene-d5 (surr.)
2-Nitrophenol
4-Nitrophenol
N-Nitroso-di-n-butylamine
N-Nitrosodimethylamine"
N-Nitrosodiphenylaminea
N-Nitroso-di-n-propylamine
N-Nitrosopiperidine
Pentachl orobenzene
Pentachloronitrobenzene
Pentachl orophenol
Perylene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenol
Phenol -d6 (surr.)
2-Picoline
Pronamide
Pyrene
Terphenyl-du (surr.)
1,2,4 , 5-Tetrachl orobenzene
2,3,4 , 6-Tetrachl orophenol
42.7
11.9
--
--
--
--
--
--
12.1
--
--
--
--
--
--
11.1
--
6.5
20.3
--
--
20.5
--
._
--
--
17.5
--
--
22.8
--
8.0
--
--
--
27.3
--
--
--
3.7
2.2
--
--
--
--
--
--
1.6
--
--
--
--
--
--
1.9
--
3.6
2.4
--
--
1.9
--
--
--
--
3.6
--
--
5.4
--
1.5
--
--

1.9
--
--
--
276
82
227
268
80
142
108
108
128
136
143
143
65
138
138
77
82
139
139
84
42
169
70
42
250
295
266
264
108
178
188
94
99
93
173
202
244
216
232
138,
95,
228
253,
79,
141
107,
107,
129,
68
115,
115,
92,
108,
108,
123,
128,
109,
109,
57,
74,
168,
130,
114,
252,
237,
264,
260,
109,
179,
94,
65,
42,
66,
175,
200,
122,
214,
230,
227
138

267
65

79
79
127

116
116
138
92
92
65
54
65
65
41
44
167
42
55
248
142
268
265
179
176
80
66
71
92
145
203
212
218
131
                                   8250A -  24
                                   Revision 1
                               September 1994

-------
TABLE 1.
(Continued)


Compound
Toxapheneb
2,4,6-Tribromophenol (surr.)
1 , 2 , 4-Tri chl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol

Retention
Time (min)
25-34
--
11.6
--
11.8
Method
Detection
Limit (jug/L)
_ _.
--
1.9
--
2.7

Primary
Ion
159
330
180
196
196

Secondary
Ion(s)
231, 233
332, 141
182, 145
198, 200
198, 200
aSee Sec.  1.3
bThese compounds are mixtures of various isomers.
(I.S.)  = Internal Standard
(surr). = Surrogate
                                   TABLE 2.
             DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQL)
                             FOR VARIOUS MATRICES8
Matrix
  Factor
Ground water                                                           10
Low-concentration soil by ultrasonic extraction with GPC cleanup      670
High-concentration soil and sludges by ultrasonic extraction       10,000
Non-water miscible waste                                          100,000
    EQL  =  [Method detection limit (see Table 1)]  X  [Factor  found  in this table].
    For  non-aqueous  samples,  the factor is on a wet-weight basis.  Sample EQLs
    are  highly matrix-dependent.  The  EQLs  to be determined herein are provided
    for  guidance  and  may  not  always  be achievable.
                                  8250A - 25
    Revision 1
September 1994

-------
                                   TABLE 3.
                  DFTPP KEY IONS AND ION ABUNDANCE CRITERIA8
       Mass             Ion Abundance Criteria
       51               30-60% of mass 198
       68               < 2% of mass 69
       70               < 2% of mass 69
      127               40-60% of mass 198
      197               < 1% of mass 198
      198               Base peak, 100% relative abundance
      199               5-9% of mass 198
      275               10-30% of mass 198
      365               > 1% of mass 198
      441               Present but less than mass 443
      442               > 40% of mass 198
      443               17-23% of mass 442
"See  Reference 3,
                                  8250A - 26                        Revision 1
                                                                September 1994

-------
                             TABLE 4.
                    CALIBRATION CHECK COMPOUNDS
Base/Neutral Fraction                   Acid Fraction
Acenaphthene                            4-Chloro-3-methyl phenol
1,4-Dichlorobenzene                     2,4-Dichlorophenol
Hexachlorobutadiene                     2-Nitrophenol
N-Nitroso-di-n-phenylamine              Phenol
Di-n-octyl phthalate                    Pentachlorophenol
Benzo(a)pyrene                          2,4,6-Tri chlorophenol
Fluoranthene
                            8250A -  27                        Revision 1
                                                          September 1994

-------
                                   TABLE 5.
          SEMIVOLATILE  INTERNAL  STANDARDS WITH  CORRESPONDING ANALYTES
                           ASSIGNED  FOR  QUANTITATION
l,4-Dichlorobenzene-D4
Naphthalene-dg
Acenaphthene-di
Aniline
Benzyl alcohol
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl)ether
2-Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl methanesulfonate
2-Fluorophenol (surr.)
Hexachloroethane
Methyl methanesulfonate
2-Methylphenol
4-Methylphenol
N-Nitrosodimethyl amine
N-Nitroso-di-n-propylamine
Phenol
Phenol-de (surr.)
2-Picoline
Acetophenone
Benzoic acid
Bis(2-chloroethoxy)methane
4-Chloroaniline
4-Chloro-3-methyl phenol
2,4-Dichlorophenol
2,6-Dichlorophenol
a,a-Dimethylphenethylamine
2,4-Dimethylphenol
Hexachlorobutadiene
Isophorone
2-Methylnaphthalene
Naphthalene
Nitrobenzene
Nitrobenzene-d8 (surr.)
2-Nitrophenol
N-Nitroso-di-n-butyl amine
N-Nitrosopiperi dine
1,2,4-Trichlorobenzene
Acenaphthene
Acenaphthylene
1-Chloronaphthalene
2-Chloronaphthalene
4-Chlorophenyl
  phenyl  ether
Dibenzofuran
Diethyl phthalate
Dimethyl  phthalate
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Fluorene
2-Fluorobiphenyl
  (surr.)
Hexachlorocyclo-
  pentadiene
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
4-Nitrophenol
Pentachlorobenzene
1,2,4,5-Tetrachloro-
  benzene
2,3,4,6-Tetrachloro-
  phenol
2,4,6-Tribromophenol
  (Surr.)
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
(surr.) = surrogate
                                  8250A - 28
                                        Revision  1
                                    September 1994

-------
                                   TABLE 5.
          SEMIVOLATILE INTERNAL STANDARDS WITH CORRESPONDING ANALYTES
                           ASSIGNED FOR QUANTITATION
                                  (Continued)
Phenanthrene-d
              10
Chrysene-d12
Perylene-d12
4-Aminobiphenyl
Anthracene
4-Bromophenyl phenyl ether
Di-n-butyl phthalate
4,6-Dinitro-2-methylphenol
Diphenylamine
1,2-Diphenylhydrazine
Fluoranthene
Hexachlorobenzene
N-Nitrosodiphenylamine
Pentachlorophenol
Pentachloronitrobenzene
Phenacetin
Phenanthrene
Pronarm'de
Benzidine
Benzo(a)anthracene
Bis(2-ethylhexyl) phthalate
Butyl benzyl phthalate
Chrysene
3,3'-Dichlorobenzidine
p-Dimethylaminoazobenzene
Pyrene
Terphenyl-d14  (surr.)
Benzo(b)fluoranthene
Benzo(k)f1uoranthene
Benzo(g,h,i)perylene
Benzo(a)pyrene
Dibenz(a,j)acridine
Dibenz(a,h)anthracene
7,12-Dimethylbenz-
  (a)anthracene
Di-n-octyl phthalate
Indeno(l,2,3-cd)pyrene
3-Methylcholanthrene
(surr.) = surrogate
                                  8250A - 29
                                        Revision 1
                                    September 1994

-------
       TABLE 6.
QC ACCEPTANCE CRITERIA8
Test
cone.
Compound (M9/L)
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benzo(a)anthracene
Benzo(b)fl uoranthene
Benzo(k)fl uoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Butyl benzyl phthalate
6-BHC
5-BHC
Bis(2-chloroethyl) ether
Bi s ( 2 -chl oroethoxy)methane
Bis(Z-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chl oronaphthal ene
4-Chlorophenyl phenyl ether
Chrysene
4, 4' -ODD
4,4'-DDE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 ,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fl uoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(M9/L)
27.6
40.2
39.0
32.0
27.6
38.8
32.3
39.0
58.9
23.4
31.5
21.6
55.0
34.5
46.3
41.1
23.0
13.0
33.4
48.3
31.0
32.0
61.6
70.0
16.7
30.9
41.7
32.1
71.4
30.7
26.5
23.2
21.8
29.6
31.4
16.7
32.5
32.8
20.7
37.2
54.7
24.9
26.3
Range
for x
(M/L)
60.1-132.3
53.5-126.0
7.2-152.2
43.4-118.0
41.8-133.0
42.0-140.4
25.2-145.7
31.7-148.0
D-195.0
D-139.9
41.5-130.6
D-100.0
42.9-126.0
49.2-164.7
62.8-138.6
28.9-136.8
64.9-114.4
64.5-113.5
38.4-144.7
44.1-139.9
D-134.5
19.2-119.7
D-170.6
D-199.7
8.4-111.0
48.6-112.0
16.7-153.9
37.3-105.7
8.2-212.5
44.3-119.3
D-100.0
D-100.0
47.5-126.9
68.1-136.7
18.6-131.8
D-103.5
D-188.8
42.9-121.3
71.6-108.4
D-172.2
70.9-109.4
7.8-141.5
37.8-102.2
Range
P> Ps
(%)
47-145
33-145
D-166
27-133
33-143
24-159
11-162
17-163
D-219
D-152
24-149
D-110
12-158
33-184
36-166
8-158
53-127
60-118
25-158
17-168
D-145
4-136
D-203
D-227
1-118
32-129
D-172
20-124
D-262
29-136
D-114
D-112
39-139
50-158
4-146
D-107
D-209
26-137
59-121
D-192
26-155
D-152
24-116
       8250A -  30
    Revision 1
September 1994

-------
                             TABLE 6.
                      QC ACCEPTANCE CRITERIA8
                            (Continued)


Compound
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitroso-di-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1 , 2 , 4-Tri chl orobenzene
4-Chloro-3-methyl phenol
2-Chlorophenol
2,4-Chlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
s = Standard deviati
x = Average recovery
p, ps = Percent recovery
D = Detected; result
a r\*4 -faln-i •> •Pixnm A f\ TCD D
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
(M9/L)
24.5
44.6
63.3
30.1
39.3
55.4
54.2
20.6
25.2
28.1
37.2
28.7
26.4
26.1
49.8
93.2
35.2
47.2
48.9
22.6
31.7
on of four recovery
for four
measured
must be
I-../.+ IOC 4
recovery
•
Range
for x
(M9/L)
55.2-100.0
D-150.9
46.6-180.2
35.6-119.6
54.3-157.6
13.6-197.9
19.3-121.0
65.2-108.7
69.6-100.0
57.3-129.2
40.8-127.9
36.2-120.4
52.5-121.7
41.8-109.0
D-172.9
53.0-100.0
45.0-166.7
13.0-106.5
38.1-151.8
16.6-100.0
52.4-129.2
measurements,
measurements,

Range
P» Ps
(%)
40-113
D-171
21-196
21-133
35-180
D-230
D-164
54-120
52-115
44-142
22-147
23-134
39-135
32-119
0-191
D-181
29-182
D-132
14-176
5-112
37-144
in M9/L-
in /ng/L.

greater than zero.
?nv. Mn+Un/J
COC T l«» r\ *•. *•» <- <
kA ^ •t' /*\ M •« •% *\ t/»r\ !•*•«** St sJ
directly on the method performance  data  in Table 7.  Where necessary, the
limits for  recovery  have  been broadened to assure  applicability of the
limits to concentrations below those used to develop Table 7.
                            8250A - 31
    Revision 1
September 1994

-------
                          TABLE 7.
METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION8
Parameter
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benzo(a)anthracene
Chloroethane
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Butyl benzyl phthalate
B-BHC
tJ-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4,4'-DOE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 , 4-Di chl orobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fluoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachl orobenzene
Hexachlorobutadiene
Hexachloroethane
Accuracy, as
recovery, x'
(M9/L)
0.96C+0.19
0.89C+0.74
0.78C+1.66
0.80C+0.68
0.88C-0.60
0.99C-1.53
0.93C-1.80
0.87C-1.56
0.90C-0.13
0.98C-0.86
0.66C-1.68
0.87C-0.94
0.29C-1.09
0.86C-1.54
1.12C-5.04
1.03C-2.31
0.84C-1.18
0.91C-1.34
0.89C+0.01
0.91C+0.53
0.93C-1.00
0.56C-0.40
0.70C-0.54
0.79C-3.28
0.88C+4.72
0.59C+0.71
0.80C+0.28
0.86C-0.70
0.73C-1.47
1.23C-12.65
0.82C-0.16
0.43C+1.00
0.20C+1.03
0.92C-4.81
1.06C-3.60
0.76C-0.79
0.39C+0.41
0.76C-3.86
0.81C+1.10
0.90C-0.00
0.87C-2.97
0.92C-1.87
0.74C+0.66
0.71C-1.01
0.73C-0.83
Single analyst
precision, s/
(M9/L)
0.15X-0.12
0.24X-1.06
0.27X-1.28
0.21X-0.32
O.lSx+0.93
0.14X-0.13
0.22X+0.43
0.19X+1.03
0.22X+0.48
0.29x+2.40
O.lSx+0.94
0.20X-0.58
0.34X+0.86
0.35x~0.99
O.lSx+1.34
0.24X+0.28
0.26x+0.73
0.13X+0.66
0.07X+0.52
0.20x-0.94
0.28x+0,13
0.29x-0.32
0.26X-1.17
0.42X+0.19
0.30X+8.51
0.13X+1.16
0.20X+0.47
0.25X+0.68
0.24X+0.23
0.28X+7.33
0.20X-0.16
0.28X+1.44
0.54x+0'.19
0.12x+1.06
0.14X+1.26
O.Zlx+1.19
0.12X+2.47
O.lSx+3.91
0.22X-0.73
0.12X+0.26
0.24X-0.56
0.33x-0.46
0.18X-0.10
0.19X+0.92
0.17X+0.67
Overall
precision,
S' (/ig/L)
0.21X-0.67
0.26X-0.54
0.43X+1.13
0.27X-0.64
0.26X-0.21
0.17X-0.28
0.29X+0.96
0.35X+0.40
0.32X+1.35
0.51X-0.44
0.53X+0.92
0.30X+1.94
0.93X-0.17
0.35X+0.10
0.26X+2.01
0.25X+1.04
0.36X+0.67
0.16X+0.66
0.13X+0.34
0.30X-0.46
0.33X-0.09
0.66X-0.96
0.39X-1.04
0.65X-0.58
0.59X+0.25
0.39X+0.60
0.24X+0.39
0.41X+0.11
0.29X+0.36
0.47X+3.45
0.26X-0.07
0.52X+0.22
1.05X-0.92
O.Zlx+1.50
0.19X+0.35
0.37X+1.19
0.63X-1.03
0.73X-0.62
0.28X-0.60
0.13X+0.61
O.BOx-0.23
0.28X+0.64
0.43X-0.52
0.26X-I-0.49
0.17X+0.80
                         8250A - 32
    Revision 1
September 1994

-------
                                   TABLE 7.
         METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION3
                                  (Continued)
Parameter
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitroso-di-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4-Chl oro-3 -methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
Accuracy, as
recovery, x'
(Mg/L)
0.78C-3.10
1.12C+1.41
0.76C+1.58
1.09C-3.05
1.12C-6.22
0.81C-10.86
0.87C+0.06
0.84C-0.16
0.94C-0.79
0.84C+0.35
0.78C+0.29
0.87C-0.13
0.71C+4.41
0.81C-18.04
1.04C-28.04
0.07C-1.15
0.61C-1.22
0.93C+1.99
0.43C+1.26
0.91C-0.18
Single analyst
precision, s/
(M9/L)
0.29x+1.46
0.27x+0.77
0.21X-0.41
0.19X+0.92
0.27x+0.68
0.35x+3.61
0.12X+0.57
0.16X+0.06
O.lSx+0.85
0.23X+0.75
O.lSx+1.46
O.lSx+1.25
O.lGx+1.21
0.38X+2.36
O.lOx+42.29
0.16X+1.94
0.38X+2.57
0.24X+3.03
0.26X+0.73
0.16X+2.22
Overall
precision,
S' (Mg/L)
O.BOx-0.44
0.33X+0.26
O.SOx-0.68
0.27X+0.21
0.44X+0.47
0.43X+1.82
0.15X+0.25
0.15X+0.31
0.21X+0.39
0.29X+1.31
0.28X+0.97
0.21X+1.28
0.22X+1.31
0.42X+26.29
0.26X+23.10
0.27x+2.60
0.44X+3.24
0.30X+4.33
0.35X+0.58
0.22X+1.81
X'
S'


C

x
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C,  in
Expected single  analyst  standard deviation of measurements  at an
average concentration of x, in p.g/1.

Expected interlaboratory standard deviation of measurements  at an
average concentration found of x, in

True value for the concentration, in

Average recovery  found  for measurements  of samples  containing  a
concentration of C, in
                                  8250A - 33
                                                        Revision 1
                                                    September 1994

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR WATER AND SOIL/SEDIMENT SAMPLES
                                    Low/Medium            Low/Medium
Surrogate Compound                     Water             Soil/Sediment
Nitrobenzene-d5                       35-114                23-120
2-Fluorobiphenyl                      43-116                30-115
Terphenyl-du                          33-141                18-137

Phenol-d6                             10-94                 24-113
2-Fluorophenol                        21-100                25-121
2,4,6-Tribromophenol                  10-123                19-122
                                  8250A - 34                        Revision 1
                                                                September 1994

-------
                                        METHOD  8250A
SEMIVOLATILE  ORGANIC  COMPOUNDS  BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS)
                 7.1 Prepare sample
                 using Method 3540,
                   3541, or 35SO
7.1 Prepare tampl*
using Method 3610
    or 3520
                                            7.1 Prepare (ample
                                            using Method 3540,
                                           3541, 3550, or 3580
                                               7.2 Cleenup
                                                 extract
                                                   7.3
                                               Recommended
                                                 GC/MS
                                                 operating
                                                conditions.
                                                  7.4
                                                  Initial
                                                Calibration.
                                                 7.5 Daily
                                              calibration - Tune
                                             GC/MS with TFTPP
                                             and check SPCC &
                                                  CCC.
                                         8250A -  35
          Revision 1
      September  1994

-------
        METHOD 8250A
          continued
 7.6.1 Screen extract
in GC/FID or GC/PID to
  eliminate too high
   concentration*.
     7.6.2 Spike
     sample with
       internal
      standard.
     7.6.3 Analyze
   extract by GC/MS
  using recommended
  column and operating
      condition*.
        7.6.4
        Doe*
   response exceed
   initial calibration
        curve
        range?
7.6.4 Dilute
  extract.
     7.7.1  Identify
     compound* by
    comparing sample
    retention time and
  sample mass spectra
     to standard*.
         7.7.2
       Quantitate
      sample* using
      internal  ltd.
       technique.
7.7.2.4 Report
results.
1
r
     C   Stop  J
         8250A -  36
                  Revision  1
             September  1994

-------
                                 METHOD 8260A

  VOLATILE  ORGANIC  COMPOUNDS  BY  GAS  CHROMATOGRAPHY/MASS  SPECTROMETRY  (GC/MS):
                          CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8260  is  used to determine volatile organic  compounds  in a
variety of solid waste matrices.  This method is applicable to nearly all types
of samples,  regardless of water content, including ground water, aqueous sludges,
caustic  liquors,  acid liquors,  waste  solvents,  oily wastes
fibrous  wastes,  polymeric  emulsions,   filter  cakes,  spent
catalysts, soils, and sediments.  The following compounds can
this method:
                          mousses,  tars,
                          carbons,   spent
                         be  determined by
Analyte
CAS No.1
  Appropriate Technique
                  Direct
Purge-and-Trap    Injection
Acetone
Acetonitrile
Acrolein (Propenal)
Acrylonitrile
Ally! alcohol
Allyl chloride
Benzene
Benzyl chloride
Bromoacetone
Bromochloromethane (I.S.)
Bromodichloromethane
4-Bromofluorobenzene (surr.)
Bromoform
Bromomethane
n-Butanol
2-Butanone (MEK)
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chlorobenzene
2 - Chi oro- 1,3 -butadiene
Chi orod i bromomethane
Chloroethane
2-Chloroethanol
bis-(2-Chloroethyl) sulfide
2-Chloroethyl vinyl ether
Chloroform
Chloromethane
Chloroprene
67-64-1
75-05-8
107-02-8
107-13-1
107-18-6
107-05-1
71-43-2
100-44-7
598-31-2
74-97-5
75-27-4
460-00-4
75-25-2
74-83-9
71-36-3
78-93^3
75-15-0
56-23-5
302-17-0
108-90-7
126-99-8
124-48-1
75-00-3
107-07-3
505-60-2
110-75-8
67-66-3
74-87-3
126-99-8
PP
PP
PP
PP
ht
a
a
a
PP
a
a
a
a
a
ht
PP
PP
a
PP
a
a
a
a
PP
PP
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
                                   8260A  -  1
                               Revision  1
                           September  1994

-------
   Appropriate  Technique
Analyte
3-Chloropropene
3-Chl oropropi oni tri 1 e
l,2-Dibromo-3-chloropropane
1,2-Dibromoethane
Dibromomethane
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
cis-l,4-Dichloro-2-butene
trans- l,4-Dichloro-2-butene
Di chl orod i f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
trans- 1 ,2-Di chl oroethene
1,2-Dichloropropane
l,3-Dichloro-2-propanol
cis-l,3-Dichloropropene
trans-l,3-Dichloropropene
1,2,3,4-Diepoxybutane
Di ethyl ether
1,4-Difluorobenzene (I.S.)
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl acetate
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
Hexachlorobutadiene
Hexachloroethane
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Isopropyl benzene
Malononitrile
Methacrylonitrile
Methanol
Methylene chloride (DCM)
Methyl methacrylate
4-Methyl-2-pentanone (MIBK)
Naphthalene
Nitrobenzene
2-Nitropropane
CAS No."
107-05-1
542-76-7
96-12-8
106-93-4
74-95-3
95-50-1
541-73-1
106-46-7
1476-11-5
110-57-6
75-71-8
75-34-3
107-06-2
75-35-4
156-60-5
78-87-5
96-23-1
10061-01-5
10061-02-6
1464-53-5
60-29-7
540-36-3
123-91-1
106-89-8
64-17-5
141-78-6
100-41-4
75-21-8
97-63-2
87-68-3
67-72-1
591-78-6
78-97-7
74-88-4
78-83-1
98-82-8
109-77-3
126-98-7
67-56-1
75-09-2
80-62-6
108-10-1
91-20-3
98-95-3
79-46-9
Purge-and-Trap
a
i
PP
a
a
a
a
a
a
PP
a
a
a
a
a
a
PP
a
a
a
a
a
PP
i
i
i
a
PP
a
a
i
PP
i
a
PP
a
PP
PP
i
a
a
PP
a
a
a
Direct
Injection
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
8260A - 2
    Revision 1
September 1994

-------
Appropriate Technique

Analyte CAS No.b
Pentachloroethane 76-01-7
2-Picoline 109-06-8
Propargyl alcohol 107-19-7
B-Propiolactone 57-57-8
Propionitrile (ethyl cyanide) 107-12-0
n-Propylamine 107-10-8
Pyridine 110-86-1
Styrene 100-42-5
1,1,1 , 2-Tetrachl oroethane 630-20-6
1,1,2,2-Tetrachloroethane 79-34-5
Tetrachloroethene 127-18-4
Toluene 108-88-3
1,2,4-Trichlorobenzene 120-82-1
1,1,1-Trichloroethane 71-55-6
1,1,2-Trichloroethane 79-00-5
Trichloroethene 79-01-6
Trichlorofluoromethane 75-69-4
1,2,3-Trichloropropane 96-18-4
Vinyl acetate 108-05-4
Vinyl chloride 75-01-4
o-Xylene 95-47-6
m-Xylene 108-38-3
p-Xylene 106-42-3
a Adequate response by this technique.
b Chemical Abstract Services Registry Number
ht Method analyte only when purged at 80°C
i Inappropriate technique for this analyte.
pc Poor chromatographic behavior.
pp Poor purging efficiency resulting in high
surr Surrogate
I.S. Internal Standard

Purge-and-Trap
i
PP
PP
PP
ht
a
i
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a

.



EQLs.


Direct
Injection
a
a
a
a
pc
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a
a








      1.2   Method 8260 can be used to quantitate most volatile organic compounds
that have boiling points below 200°C and that are insoluble or slightly soluble
in water.  Volatile water-soluble compounds can be included in this analytical
technique.  However,  for the more soluble compounds,  quantitation limits are
approximately  ten  times  higher because  of poor  purging  efficiency.    Such
compounds  include  low-molecular-weight  halogenated hydrocarbons,  aromatics,
ketones, nitriles,  acetates,  acrylates,  ethers,  and  sulfides.  See Tables  1 and
2 for lists of analytes and retention times that have been evaluated on a purge-
                                   8260A  - 3                         Revision 1
                                                                September 1994

-------
and-trap  GC/MS  system.   Also,  the method detection  limits for 25  ml sample
volumes are presented.  The following  analytes  are also amenable to analysis by
Method 8260:

      Bromobenzene                  1-Chlorohexane
      n-Butylbenzene                2-Chlorotoluene
      sec-Butyl benzene              4-Chlorotoluene
      tert-Butylbenzene             Crotonaldehyde
      Chioroaceton i tri1e            Di bromofluoromethane
      1-Chlorobutane                cis-l,2-Dichloroethene
      1,3-Dichloropropane           Methyl-t-butyl ether
      2,2-Dichloropropane           Pentafluorobenzene
      1,1-Dichloropropene           n-Propylbenzene
      Fluorobenzene                 1,2,3-Trichlorobenzene
      p-1sopropyltoluene            1,2,4-Trimethylbenzene
      Methyl acrylate               1,3,5-Trimethylbenzene

      1.3   The  estimated  quantitation  limit   (EQL)  of  Method 8260  for  an
individual compound is somewhat instrument dependent.  Using  standard quadrupole
instrumentation,  limits should  be  approximately  5  jugAg  (wet  weight)  for
soil/sediment samples, 0.5  mg/kg (wet  weight) for wastes, and 5 /xg/L for ground
water (see Table 3).  Somewhat  lower  limits may be  achieved using  an ion trap
mass spectrometer or other  instrumentation of improved design.  No matter which
instrument is used,  EQLs will  be  proportionately higher for  sample extracts and
samples that require dilution  or reduced sample size to avoid saturation of the
detector.

      1.4   Method 8260  is based upon a purge-and-trap, gas chromatographic/mass
spectrometric (GC/MS) procedure.  This method is restricted  to use by, or under
the supervision of, analysts experienced in the  use of purge-and-trap systems and
gas chromatograph/mass spectrometers,  and  skilled in the interpretation of mass
spectra and their use as a quantitative tool.

      1.5   An  additional method for  sample introduction  is direct injection.
This technique  has  been tested for  the  analysis  of waste oil diluted with
hexadecane  1:1  (vol/vol) and  may  have application for  the analysis  of some
alcohols and aldehydes in aqueous samples.


2.0   SUMMARY OF METHOD

      2.1   The  volatile compounds  are introduced into the gas chromatograph by
the  purge-and-trap  method  or  by  direct   injection (in  limited  applications).
Purged  sample  components are  trapped in  a  tube containing  suitable  sorbent
materials.  When purging is complete, the sorbent tube  is heated and backflushed
with helium  to  desorb trapped  sample components.   The analytes are desorbed
directly to  a  large  bore  capillary or cryofocussed on a  capillary  precolumn
before being flash  evaporated to  a  narrow bore capillary  for  analysis.   The
column is temperature programmed to separate the analytes which are then detected
with a mass spectrometer (MS)  interfaced  to the gas chromatograph.   Wide bore
capillary columns require a jet separator,  whereas narrow bore capillary columns
can be directly interfaced to the ion source.
                                   8260A -  4                         Revision 1
                                                                September 1994

-------
      2.2   If the above sample introduction techniques  are  not applicable,  a
portion of the sample is dispersed in solvent to dissolve the volatile organic
constituents.  A portion of the solution is combined with organic-free reagent
water  in  the  purge  chamber.   It  is  then  analyzed  by  purge-and-trap  GC/MS
following the normal  water method.

      2.3   Analytes  eluted from the capillary  column  are introduced into the
mass spectrometer via a  jet separator or a direct connection.  Identification of
target analytes is accomplished by comparing their mass spectra with the electron
impact (or electron impact-like) spectra of authentic standards.  Quantitation
is accomplished by comparing the response of a major  (quantitation) ion relative
to an internal standard with a five-point calibration curve.

      2.4   The method includes specific calibration and quality control  steps
that replace the general requirements in Method 8000.


3.0   INTERFERENCES

      3.1   Major contaminant  sources  are volatile materials  in the laboratory
and impurities in the  inert purging gas and  in the sorbent trap.   The use of non-
polytetrafluoroethylene  (PTFE)  thread  sealants,  plastic  tubing,   or  flow
controllers with rubber components should be avoided since such materials out-gas
organic compounds  which will  be concentrated  in the  trap  during  the  purge
operation.  Analyses  of  calibration and reagent  blanks provide information about
the presence  of  contaminants.   When potential   interfering peaks  are noted  in
blanks, the  analyst   should  change the  purge  gas  source and  regenerate  the
molecular sieve  purge gas  filter  (Figure  1).   Subtracting  blank  values  from
sample results is not permitted.  If reporting  values not corrected for blanks
result in  what the laboratory feels is  a  false positive for a sample, this should
be fully explained in text accompanying the uncorrected data.

      3.2   Interfering  contamination  may  occur  when  a  sample  containing  low
concentrations of volatile  organic compounds is  analyzed  immediately  after  a
sample  containing  high  concentrations  of  volatile organic  compounds.    The
preventive technique  is rinsing of the purging apparatus and sample syringes with
two portions of organic-free reagent water between samples.   After analysis  of
a sample containing high concentrations  of volatile  organic  compounds,  one  or
more calibration blanks  should be analyzed to check for cross contamination.  For
samples containing large amounts of water soluble materials,  suspended solids,
high boiling compounds or high  concentrations of compounds being determined,  it
may be necessary to wash the  purging device with a soap solution, rinse it with
organic-free reagent  water,  and then dry the purging device in an oven at 105°C.
In extreme situations, the whole purge and trap device may require dismantling
and cleaning.  Screening of the samples prior to purge and trap GC/MS analysis
is highly  recommended  to prevent contamination of the system.   This is especially
true for soil  and waste samples. Screening may be accomplished with an automated
headspace technique or  by Method  3820  (Hexadecane Extraction and  Screening  of
Purgeable Organics).

            3.2.1  The low  purging efficiency of  many analytes from a 25  ml
      sample often results in significant concentrations remaining  in  the sample
      purge vessel after analysis.  After removal of  the analyzed sample aliquot


                                   8260A -  5                         Revision  1
                                                                September  1994

-------
      and  three  rinses of  the purge  vessel  with analyte  free  water,  it  is
      required that the empty vessel  be subjected  to a heated purge cycle prior
      to  the  analysis of another  sample in the  same  purge vessel  to reduce
      sample to sample carryover.

      3.3   Special  precautions must be taken to analyze for methylene chloride.
The analytical and sample storage area should be isolated from all  atmospheric
sources of methylene chloride.  Otherwise random background levels will result.
Since  methylene   chloride   will   permeate  through  PTFE   tubing,   all   gas
chromatography carrier gas  lines and  purge  gas plumbing  should be  constructed
from stainless steel or copper tubing.   Laboratory clothing worn by the analyst
should be clean since  clothing previously  exposed  to methylene chloride fumes
during   liquid/liquid  extraction   procedures  can   contribute  to   sample
contamination.

      3.4   Samples  can  be  contaminated by  diffusion  of  volatile  organics
(particularly methylene chloride and fluorocarbons) through the  septum seal into
the sample during  shipment and storage.  A trip  blank prepared from organic-free
reagent water and carried through the sampling and handling  protocol  can serve
as a check on such contamination.

      3.5   Use of sensitive mass spectrometers to  achieve lower detection level
will increase the  potential  to  detect laboratory contaminants as interferences.

      3.6   Direct injection - Some  contamination may be  eliminated by baking out
the column  between analyses.   Changing the  injector liner will  reduce  the
potential for cross-contamination.   A portion of the analytical column may need
to be  removed  in the case of extreme contamination.  Use  of direct  injection will
result in the need for more frequent instrument maintenance.

      3.7   If hexadecane is added to samples or petroleum samples are analyzed,
some  chromatographic  peaks  will  elute after  the  target analytes.   The  oven
temperature program must include  a  post-analysis bake out  period  to ensure that
semi -volatile hydrocarbons are volatilized.


4.0   APPARATUS AND MATERIALS

      4.1   Purge-and-trap  device - aqueous  samples, described  in Method 5030.

      4.2   Purge-and-trap device - solid samples,  described  in Method  5030.

      4.3   Injection  port liners (HP catalogue #18740-80200,  or equivalent) are
modified  for  direct  injection analysis  by  placing  a  1-cm plug  of pyrex wool
approximately 50-60 mm down  the length
of  the  injection  port  towards  the
oven.  An 0.53 mm ID column is mounted    s«Pt«.m
1 cm into the liner  from the oven side
of  the  injection port,  according  to
manufacturer's specifications.                       Modified Injector
                                   8260A -  6                         Revision 1
                                                                September 1994

-------
4.4   Gas chromatography/mass spectrometer/data system

      4.4.1  Gas  chromatograph  -  An  analytical  system complete  with a
temperature-programmable   gas   chromatograph   suitable   for  splitless
injection or interface to purge-and-trap apparatus.  The system includes
all  required  accessories,  including  syringes,  analytical  columns,   and
gases.  The  GC  should be equipped with variable  constant differential flow
controllers so that the column flow rate will remain constant throughout
desorption  and   temperature  program  operation.     For   some  column
configurations,  the  column  oven must be cooled  to < 30°C,  therefore, a
subambient oven controller may be required.  The capillary column should
be directly coupled to the source.

            4.4.1.1      Capillary precolumn interface when using cryogenic
      cooling  -  This  device  interfaces the  purge  and  trap concentrator to
      the capillary  gas  chromatograph.    The  interface condenses   the
      desorbed  sample components and focuses them into a narrow band on an
      uncoated  fused silica capillary precolumn.   When the  interface is
      flash  heated,  the  sample  is  transferred to  the  analytical capillary
      column.

                  4.4.1.1.1  During   the    cryofocussing    step,    the
            temperature of the  fused silica in the interface is maintained
            at  -150°C  under a  stream of  liquid  nitrogen.   After  the
            desorption period,  the  interface  must be capable  of rapid
            heating  to  250°C  in  15  seconds  or  less  to  complete  the
            transfer of analytes.

      4.4.2  Gas chromatographic columns

            4.4.2.1      Column 1  -  60 m  x 0.75 mm  ID capillary column
      coated with  VOCOL  (Supelco), 1.5 jttm film thickness, or equivalent.

            4.4.2.2      Column 2  - 30 - 75  m x 0.53 mm ID capillary column
      coated  with DB-624  (J&W  Scientific),  Rtx-502.2 (RESTEK),  or VOCOL
      (Supelco),  3 jLtm film thickness,  or  equivalent.

            4.4.2.3      Column 3  -  30 m  x 0.25 - 0.32  mm  ID capillary
      column  coated with 95% dimethyl - 5% diphenyl  polysiloxane (DB-5,
      Rtx-5, SPB-5, or equivalent), 1 jitm film thickness.

            4.4.2.4      Column 4  -  60 m  x 0.32 mm  ID capillary column
      coated  with  DB-624  (J&W  Scientific),  1.8 jum  film thickness,  or
      equivalent.

      4.4.3  Mass spectrometer  - Capable of scanning from 35  to  300  amu
every 2  sec  or  less,  using  70 volts  (nominal)  electron energy  in  the
electron impact ionization mode. The mass spectrometer  must be capable of
producing a mass spectrum for p-Bromofluorobenzene (BFB)  which meets  all
of the criteria in Table 4 when 5-50  ng of the GC/MS tuning standard (BFB)
is  injected  through the  GC.    To ensure  sufficient precision  of  mass
spectral data,  the desirable MS scan rate allows acquisition of at least
five spectra while a sample component elutes from the GC.


                             8260A - 7                         Revision 1
                                                          September 1994

-------
                  4.4.3.1      The ion trap mass spectrometer may be used if it
            is capable of axial modulation to reduce ion-molecule reactions and
            can  produce  electron impact-like  spectra  that match those  in the
            EPA/NIST  Library.   In an ion trap mass  spectrometer,  because ion-
            molecule reactions with water and methanol may produce interferences
            that coelute with chloromethane and chloroethane, the base peak for
            both of these analytes will  be  at  m/z  49.   This  ion  should be used
            as the quantitation ion in this  case.  The mass spectrometer must be
            capable of producing  a mass spectrum for BFB which meets all of the
            criteria  in Table 3 when  5 or 50 ng are  introduced.

            4.4.4  GC/MS  interface -  Two  alternatives are used to interface the
      GC to the  mass  spectrometer.

                  4.4.4.1      Direct coupling by inserting the column into the
            mass spectrometer is  generally  used for  0.25-0.32 mm id columns.

                  4.4.4.2      A  separator including an all-glass transfer line
            and  glass enrichment  device or split  interface  is  used  with  an
            0.53 mm column.

                  4.4.4.3      Any enrichment device  or transfer line can be used
            if  all  of the  performance  specifications  described  in  Sec.  8
            (including acceptable calibration at 50 ng or less) can be  achieved.
            GC-to-MS  interfaces constructed entirely of glass or of glass-lined
            materials  are recommended.   Glass  can  be deactivated by silanizing
            with dichlorodimethylsilane.

            4.4.5  Data system -  A  computer system that allows  the continuous
      acquisition  and storage  on machine-readable  media  of all mass  spectra
      obtained throughout the duration of  the  chromatographic program must be
      interfaced to the mass spectrometer.  The computer must have software that
      allows searching any  GC/MS data file for  ions  of a  specified  mass and
      plotting such ion abundances  versus  time or scan number.  This  type of
      plot is defined  as an Extracted Ion Current  Profile  (EICP).  Software must
      also  be available  that allows  integrating the  abundances  in  any EICP
      between specified  time or  scan-number limits.  The most recent version of
      the EPA/NIST Mass  Spectral  Library should also be available.

      4.5   Microsyringes -  10, 25,  100,  250,  500, and  1,000  juL.

      4.6   Syringe valve - Two-way, with Luer ends (three each), if applicable
to the purging device.

      4.7   Syringes  - 5, 10, or  25  mL,  gas-tight with  shutoff valve.

      4.8   Balance -  Analytical, 0.0001  g, and top-loading,  0.1 g.

      4.9   Glass  scintillation  vials -  20  mL,  with  Teflon lined screw-caps or
glass culture tubes with Teflon lined screw-caps.

      4.10  Vials  - 2  mL, for GC  autosampler.
                                   8260A  -  8                         Revision 1
                                                                September 1994

-------
      4.11  Disposable pipets - Pasteur.

      4.12  Volumetric flasks, Class A  -  10 ml and  100  ml,  with ground-glass
stoppers.

      4.13  Spatula - Stainless steel.


5.0   REAGENTS

      5.1   Reagent grade  inorganic chemicals shall  be used in all  tests.  Unless
otherwise indicated, it is intended that all inorganic reagents shall  conform to
the  specifications of the  Committee on  Analytical  Reagents  of  the American
Chemical Society, where such specifications are available.  Other grades may be
used, provided it is first ascertained  that the reagent is of  sufficiently high
purity to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free  reagent  water  - All references to water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Methanol,  CH3OH  - Pesticide quality or equivalent, demonstrated to
be free of analytes.  Store  apart from other solvents.

      5.4   Reagent Hexadecane -  Reagent hexadecane is defined  as hexadecane in
which interference is not  observed at the method detection limit of compounds of
interest.

            5.4.1 In order to  demonstrate  that all  interfering volatiles have
      been  removed  from  the  hexadecane,  a  direct  injection  blank must  be
      analyzed.

      5.5   Polyethylene  glycol,  H(OCH2CH2)nOH  -  Free of  interferences  at the
detection limit of the target analytes.

      5.6   Hydrochloric  acid (1:1  v/v),  HC1  - Carefully  add a measured volume
of concentrated HC1 to an equal volume of organic-free reagent water.

      5.7   Stock solutions - Stock solutions may be prepared from pure standard
materials or purchased as  certified solutions.   Prepare stock standard solutions
in methanol, using assayed liquids or gases, as appropriate.

            5.7.1 Place about 9.8 ml  of methanol in a 10 ml tared ground-glass-
      stoppered volumetric flask.  Allow  the  flask to  stand,  unstoppered, for
      about  10 minutes or until all  alcohol-wetted surfaces have dried.   Weigh
      the flask to the nearest 0.0001 g.

            5.7.2 Add the assayed reference material, as  described below.

                  5.7.2.1       Liquids - Using a 100 fj.1 syringe, immediately add
            two  or  more drops of  assayed  reference  material  to the flask;  then
            reweigh.   The liquid must  fall directly into the alcohol  without
            contacting the neck of the  flask.
                                  8260A  - 9                         Revision 1
                                                                September 1994

-------
            5.7.2.2     Gases  -  To prepare standards  for  any compounds
      that boil below 30°C (e.g.  bromomethane, chloroethane, chloromethane,
      or  vinyl  chloride),  fill  a 5 ml  valved  gas-tight syringe with the
      reference  standard to the 5.0 ml mark.   Lower the needle  to  5 mm
      above the methanol meniscus.  Slowly introduce the  reference standard
      above the surface of the liquid.  The heavy gas will rapidly dissolve
      in  the rnethanol.   Standards may also be prepared  by using a lecture
      bottle  equipped  with  a  Hamilton  Lecture Bottle  Septum (#86600).
      Attach Teflon tubing to the side arm relief valve and direct a gentle
      stream of  gas  into the methanol meniscus.

      5.7.3  Reweigh,  dilute to  volume,  stopper,  and then mix by inverting
the flask several times.   Calculate the concentration  in  milligrams per
liter (mg/L)  from the net gain in weight.  When compound purity  is assayed
to  be 96% or  greater, the weight may be used  without  correction  to
calculate the  concentration of  the  stock standard.  Commercially prepared
stock standards may be  used at  any  concentration if they are certified by
the manufacturer or by an independent  source.

      5.7.4  Transfer the  stock standard solution  into a  bottle  with  a
Teflon lined screw-cap.  Store, with minimal  headspace, at  -10°C to  -20°C
and protect from light.

      5.7.5  Prepare  fresh  standards   for  gases  weekly   or   sooner  if
comparison with check standards indicates a problem.   Reactive compounds
such as 2-chloroethyl vinyl  ether and styrene may need to be prepared more
frequently.  All  other standards  must   be replaced after  six  months, or
sooner if comparison with check standards indicates a problem.  Both gas
and  liquid  standards  must  be monitored  closely  by comparison  to the
initial calibration curve  and by comparison to QC check  standards.  It may
be  necessary  to replace the standards more  frequently if either check
exceeds a 20% drift.

      5.7.6  Optionally calibration using a certified gaseous mixture can
be  accomplished  daily  utilizing commercially  available gaseous analyte
mixture  of  bromomethane,   chloromethane,  chloroethane, vinyl  chloride,
dichlorodifluoromethane  and trichlorofluoromethane  in nitrogen.  These
mixtures  of  documented quality  are  stable  for  as  long   as  six months
without refrigeration.  (VOA-CYL III, RESTEK  Corporation,  Cat. #20194 or
equivalent).

            5.7.6.1     Preparation of Calibration Standards  From a Gas
      Mixture

                  5.7.6.1.1  Before removing  the  cylinder shipping cap,
            be sure the valve is completely closed (turn clockwise).  The
            contents are  under pressure and  should  be used  in  a well-
            ventilated area.

                  5.7.6.1.2  Wrap  the pipe thread end of the Luer fitting
            with Teflon tape.   Remove the shipping cap from the cylinder
            and replace it with the Luer fitting.
                            8260A - 10                        Revision 1
                                                          September 1994

-------
       5.7.6.1.3   Transfer half the working standard containing
other  analytes,  internal standards,  and  surrogates  to the
purge  apparatus.

       5.7.6.1.4   Purge  the  Luer fitting and stem on the gas
cylinder prior to sample removal using the following sequence:

       a)     Connect either  the  100 juL or  500 juL  Luer syringe
             to the inlet fitting of the cylinder.

       b)     Make sure the on/off valve on the syringe  is  in
             the open position.

       c)     Slowly  open the   valve  on   the  cylinder  and
             withdraw a  full syringe volume.

       d)     Be sure to close the valve on the cylinder before
             you withdraw the syringe from the  Luer  fitting.

       e)     Expel   the  gas   from  the  syringe  into  a  well-
             ventilated  area.

       f)     Repeat steps a  through e one more time to  fully
             purge the fitting.

       5.7.6.1.5   Once the fitting  and stem have  been purged,
quickly withdraw  the volume of gas you require  using  steps
5.6.6.1.4(a) through  (d).   Be sure to close the  valve on the
cylinder and syringe before  you  withdraw the syringe from the
Luer fitting.

       5.7.6.1.6   Open the syringe  on/off  valve for 5 seconds
to reduce the syringe pressure to atmospheric pressure.  The
pressure in the cylinder is -30 psi.

       5.7.6.1.7   The gas mixture should be quickly transferred
into the reagent water through the female Luer fitting located
above  the purging vessel.

       NOTE:  Make   sure  the arrow  on  the 4-way valve   is
             pointing toward  the  female  Luer fitting  when
             transferring the sample from the syringe. Be sure
             to switch the  4-way  valve back  to the closed
             position before removing  the syringe  from the
             Luer  fitting.

       5.7.6.1.8   Transfer the  remaining half  of  the working
standard into the  purging vessel.  This procedure  insures that
the  total  volume  of gas mix  is  flushed into  the  purging
vessel, with none remaining in the valve or lines.

       5.7.6.1.9   Concentration   of  each  compound   in  the
cylinder is typically 0.0025
                8260A - 11                        Revision 1
                                              September 1994

-------
                        5.7.6.1.10 The following are the recommended gas volumes
                  spiked  into  5  ml  of  water  to  produce  a typical  5-point
                  calibration:

                        Gas                     Calibration
                        Volume                  Concentration
                         40 ML                      20 M9/L
                        100 ML                      50 M9/L
                        200 juL                     100 jug/L
                        300 ML                     150 M9/L
                        400 ML                     200 M9/L

                        5.7.6.1.11 The following are the recommended gas volumes
                  spiked  into  25 mL  of water  to produce  a typical  5-point
                  calibration:

                        Gas                     Calibration
                        Volume                  Concentration
                         10 ML                       1 M9/L
                         20 ML                       2 M9/L
                         50 ML                       5 M9/L
                        100 ML                      10 M9/L
                        250 ML                      25 M9/L

      5.8   Secondary  dilution  standards  -  Using  stock standard  solutions,
prepare in methanol, secondary dilution  standards  containing the  compounds of
interest, either singly  or mixed together.  Secondary dilution standards must be
stored with  minimal  headspace  and should  be  checked  frequently  for  signs of
degradation  or  evaporation,  especially  just prior  to  preparing  calibration
standards from them.  Store in a vial  with no headspace for one  week only.

      5.9   Surrogate  standards  -  The  surrogates  recommended are toluene-d8,
4-bromofluorobenzene,  l,2-dichloroethane-d4,  and dibromofluoromethane.   Other
compounds may be used as surrogates,  depending upon the analysis requirements.
A stock surrogate solution in methanol should be prepared as described above, and
a surrogate  standard spiking  solution  should be prepared from  the  stock at a
concentration of 50-250 M9/l° ml  in methanol.   Each water  sample undergoing
GC/MS analysis must be spiked with 10 ML of the surrogate  spiking solution prior
to analysis.

            5.9.1  If a  more sensitive  mass spectrometer is employed to achieve
      lower detection levels, more dilute surrogate solutions may be required.

      5.10  Internal  standards   -  The  recommended   internal   standards  are
fluorobenzene, chlorobenzene-d5, and l,4-dichlorobenzene-d4.  Other  compounds may
be used as internal standards as long as they have retention times similar  to the
compounds being detected by GC/MS. Prepare internal standard stock  and secondary
dilution  standards in methanol  using the procedures described in Sees. 5.7 and
5.8.  It  is recommended  that  the  secondary dilution standard  should be prepared
at a concentration of 25 mg/L of each  internal  standard compound.   Addition of
10 ML of this standard  to 5.0 ml  of sample  or calibration  standard would  be the
equivalent of 50 M9/L-
                                  8260A - 12                        Revision 1
                                                                September 1994

-------
            5.10.1       If  a more sensitive mass  spectrometer  is employed to
      achieve  lower  detection  levels,  more dilute internal standard solutions
      may  be  required.   Area counts of  the  internal  standard peaks should be
      between  50-200% of  the  area  of the  target  analytes  in  the mid-point
      calibration analysis.

      5.11  4-Bromofluorobenzene (BFB)  standard - A standard solution containing
25 ng//iL of BFB  in methanol  should be  prepared.

            5.11.1       If  a more sensitive mass  spectrometer  is employed to
      achieve lower detection levels, a more dilute BFB standard  solution may be
      required.

      5.12  Calibration  standards -  Calibration standards at a minimum of five
concentrations should be prepared from  the secondary dilution of  stock standards
(see Sees.  5.7 and 5.8).  Prepare these solutions in organic-free reagent water.
One of  the concentrations should be  at  a concentration near,  but above, the
method detection  limit.  The remaining concentrations should correspond to the
expected range of concentrations found  in real samples but should not exceed the
working range of  the  GC/MS system. Each standard should contain each analyte for
detection  by  this method.   It is EPA's  intent  that  all  target  analytes  for a
particular analysis be included in the  calibration  standard(s).  However, these
target analytes may  not  include  the entire List of Analytes  (Sec. 1.1) for which
the method has been  demonstrated.   However,  the laboratory shall  not report a
quantitative result  for  a target analyte that  was not included in the calibration
standard(s).  Calibration standards must be  prepared daily.

      5.13  Matrix spiking  standards  -  Matrix  spiking  standards should  be
prepared from  volatile  organic compounds which will  be  representative of the
compounds being investigated.  At a minimum, the matrix spike should include 1,1-
dichloroethene,  trichloroethene,  chlorobenzene,  toluene, and benzene.   It is
desirable  to  perform a matrix  spike  using compounds found  in  samples.   Some
permits may require  spiking specific compounds of interest, especially if they
are polar  and would  not  be  represented  by  the  above listed compounds.   The
standard should   be  prepared in  methanol,  with  each compound  present  at  a
concentration of 250 jiig/10.0 ml.

            5.13.1       If a  more sensitive mass  spectrometer  is employed to
      achieve lower detection levels, more dilute matrix spiking solutions may
      be required.

      5.14  Great care must be  taken  to maintain the integrity of all  standard
solutions.   It is recommended  all standards  in methanol  be stored at  -10°C to
-20°C  in amber bottles with Teflon lined  screw-caps.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material  to this chapter,  Organic Analytes, Sec.
4.1.
                                  8260A - 13                        Revision 1
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Three  alternate  methods  are provided for sample introduction.  All
internal standards, surrogates, and matrix spikes (when applicable) must be added
to samples before introduction.

            7.1.1  Direct  injection   -   in  very  limited  application,   (e.g.,
      volatiles in waste  oil  or  aqueous  process  wastes) direct  injection of
      aqueous  samples  or  samples  diluted  according  to Method  3585  may be
      appropriate. Direct injection has been used  for the analysis of volatiles
      in waste  oil  (diluted 1:1  with  hexadecane)  and for determining  if the
      sample is ignitable  (aqueous  injection,  Methods 1010 or  1020).   Direct
      injection is only  permitted  for  the  determination of volatiles  at the
      toxicity  characteristic  (TC)  regulatory limits,  at concentrations in
      excess of 10,000 /xg/L, or for water-soluble compounds that do not purge.

            7.1.2  Purge-and-trap  for   aqueous  samples,  see  Method  5030  for
      details.

            7.1.3  Purge-and-trap for solid samples, see Method 5030 for details.

      7.2   Recommended  Chromatographic conditions

            7.2.1  General:

                  Injector temperature:         200-225°C
                  Transfer line temperature:     250-300°C

            7.2.2  Column 1  (A  sample chromatogram  is presented  in Figure 5)

                  Carrier  gas  (He) flow rate:    15 mL/min
                  Initial  temperature:           10°C,  hold  for  5 minutes
                  Temperature  program:           6°C/min to  160°C
                  Final  temperature:            160°C,  hold until all  expected
                                                compounds have  eluted.

            7.2.3  Column  2,  Cryogenic  cooling  (A  sample  chromatogram  is
      presented in Figure 6)

                  Carrier  gas  (He) flow rate:    15 mL/min
                  Initial  temperature:           10°C,  hold  for  5 minutes
                  Temperature  program:           6°C/min to  160°C
                  Final  temperature:            160°C,  hold until all  expected
                                                compounds have  eluted.

            7.2.4  Column 2, Non-cryogenic cooling  (A sample  chromatogram is
      presented in Figure 7).  It  is recommended that carrier gas  flow and split
      and make-up gases be set  using performance of  standards as guidance.  Set
      the carrier gas head  pressure to «  10 psi and the  split  to « 30  mL/min.
      Optimize the make-up gas  flow  for the separator  (approximately 30 mL/min)
      by injecting BFB, and determining the optimum response when varying the
      make-up gas.   This  will  require  several  injections of BFB.   Next, make
      several injections of the volatile working standard with  all analytes of


                                  8260A - 14                        Revision 1
                                                                September  1994

-------
interest.  Adjust the carrier and  split  to provide optimum chromatography
and response.  This is an especially  critical  adjustment  for the volatile
gas analytes.  The head pressure should  optimize  between  8-12  psi and the
split  between  20-60 mL/min.   The use  of  the splitter  is  important to
minimize the effect  of water on analyte response,  to allow the use of a
larger volume of helium during  trap desorption, and to slow column flow.

      Initial  temperature:     45°C, hold for 2 minutes
      Temperature program:     8°C/min to 200°C
      Final  temperature:      200°C, hold for 6 minutes.

      A trap preheated  to  150°C prior to trap  desorption is required to
provide adequate chromatography of the gas analytes.

      7.2.5  Column 3 (A sample chromatogram is presented in Figure 8)

            Carrier gas (He) flow rate:    4 mL/min
            Initial temperature:          10°C, hold for 5 minutes
            Temperature program:          6°C/min to 70°C, then  15°C/min
                                          to 145°C
            Final temperature:            145°C,  hold until  all expected
                                          compounds have eluted.

      7.2.6  Direct injection - Column 2

            Carrier gas (He) flow rate:    4 mL/min
            Column:                       J&W DB-624, 70m x 0.53 mm
            Initial temperature:          40°C, hold for 3 minutes
            Temperature program:          8°C/min
            Final temperature:            260°C,  hold until  all expected
                                          compounds have eluted.
            Column Bake out (direct inj): 75 minutes
            Injector temperature:          200-225°C
            Transfer line temperature:     250-300°C

      7.2.7  Direct Split Interface -  Column 4

            Carrier gas (He) flow rate:    1.5 mL/min
            Initial temperature:          35°C, hold for 2 minutes
            Temperature program:          4°C/min to 50°C
                                          10°C/min to 220°C
            Final temperature:            220°C, hold until  all expected
                                          compounds have eluted
            Split ratio:                  100:1
            Injector temperature:          125°C

7.3   Initial  calibration  - the recommended  MS operating conditions

      Mass range:              35-260 amu
      Scan time:               0.6-2 sec/scan
      Source temperature:      According  to manufacturer's  specifications
                            8260A - 15                        Revision 1
                                                          September 1994

-------
      Ion trap only:          Set axial modulation, manifold temperature,
                             and   emission   current   to  manufacturer's
                             recommendations

      7.3.1  Each GC/MS system must  be  hardware-tuned to meet the criteria
in Table 4  for a  5-50  ng injection  or  purging  of 4-bromofluorobenzene
(2 p.L injection of the BFB standard).  Analyses must not  begin until these
criteria are met.

      7.3.2  Set up the purge-and-trap  system as outlined  in Method 5030 if
purge-and-trap  analysis  is  to  be  utilized.   A set of  at least  five
calibration  standards containing  the method analytes  is  needed.   One
calibration  standard should  contain  each  analyte  at  a  concentration
approaching but greater than the method detection limit (Table 1) for that
compound; the other  calibration  standards  should contain analytes  at
concentrations that define the range of the method.  Calibration should be
done  using  the  sample  introduction  technique  that  will  be  used  for
samples.  For  Method 5030, the  purging efficiency for  5 mL of water is
greater  than  for  25 ml.   Therefore, develop  the standard curve  with
whichever volume of  sample that will  be analyzed.

            7.3.2.1      To prepare  a  calibration standard for purge-and-
      trap or aqueous direct injection,  add an  appropriate volume  of a
      secondary dilution standard  solution to an aliquot of organic-free
      reagent  water in a volumetric flask.  Use a  microsyringe and rapidly
      inject the alcoholic standard into  the expanded area of  the filled
      volumetric flask.   Remove the needle  as quickly as  possible after
      injection.  Mix by inverting the  flask three times only.  Discard the
      contents contained in the neck of the flask.  Aqueous standards are
      not stable and  should be prepared daily.  Transfer 5.0 ml (or 25 ml
      if lower detection limits are required) of each standard to a gas
      tight  syringe along  with 10 /LtL of internal  standard.  Then transfer
      the contents  to a purging  device or  syringe.   Perform purge-and-trap
      or direct injection as  outlined  in  Method  5030.

            7.3.2.2      To prepare  a calibration  standard  for  direct
      injection analysis of oil, dilute standards in  hexadecane.

      7.3.3  Tabulate  the  area  response of the characteristic  ions  (see
Table  5) against  concentration  for each  compound  and  each  internal
standard.  Calculate  response factors (RF) for each compound relative to
one of  the  internal  standards.   The  internal standard  selected  for the
calculation of the RF for a compound should  be the  internal standard that
has a retention time closest  to the  compound being measured (Sec. 7.6.2).
The RF  is calculated  as follows:
                            8260A - 16                        Revision 1
                                                          September 1994

-------
                         RF =  (AxC,J/(AisCJ

where:

      Ax     =     Area  of  the  characteristic  ion for the compound being
                  measured.
      Ais     =     Area   of the  characteristic  ion  for  the  specific
                  internal  standard.
      Cis     =     Concentration of the specific  internal standard.
      Cx     =     Concentration of the compound  being measured.

      7.3.4 The  average RF  must  be  calculated and  recorded  for each
compound using the  five RF values  calculated  for each compound from the
initial (5-point) calibration curve.  A system performance check  should be
made before this  calibration curve  is  used.   Five  compounds (the System
Performance Check Compounds, or SPCCs) are checked for a minimum average
relative  response  factor.   These  compounds  are  chloromethane;  1,1-
dichloroethane; bromoform;  1,1,2,2-tetrachloroethane; and chlorobenzene.
These compounds are  used to check  compound  instability and to check for
degradation caused  by contaminated  lines  or  active  sites in the system.
Examples of these occurrences are:

            7.3.4.1      Chloromethane  - This  compound is the most likely
      compound to be lost if the  purge  flow  is too  fast.

            7.3.4.2      Bromoform - This compound is one  of the compounds
      most  likely  to be  purged  very poorly if  the purge flow is too slow.
      Cold  spots  and/or  active  sites in the  transfer lines may adversely
      affect  response.   Response  of the  quantitation ion  (m/z  173)  is
      directly affected by  the  tuning  of   BFB  at  ions  m/z  174/176.
      Increasing  the m/z 174/176  ratio relative to m/z 95  may improve
      bromoform response.

            7.3.4.3      Tetrachloroethane and 1,1-dichloroethane - These
      compounds are  degraded by contaminated transfer lines in purge-and-
      trap  systems and/or active  sites  in  trapping materials.

      7.3.5 Using the RFs   from the  initial   calibration,  calculate  and
record the percent relative standard deviation (%RSD) for all compounds.
The percent RSD is calculated as  follows:


                            SD
                    % RSD = — x 100%
                            RFX
where:
      RSD   =     Relative standard deviation.
      RFX    =     mean of 5 initial RFs for a compound.
      SD    =     standard deviation of the 5 initial RFs for a compound.
                            8260A - 17                        Revision 1
                                                          September 1994

-------
                   so  =
E
(RFi-RF):

  n-1
            where:

                  RFj     =  RF  for  each  of  the  5 calibration  levels
                  N      =  number  of  RF values  (i.e.,  5)

      The  percent  relative  standard deviation should be less than 15% for
each compound.  However, the  %RSD for  each individual Calibration Check
Compound  (CCC) must be less than 30%.  The CCCs are:

      1,1-Dichloroethene,
      Chloroform,
      1,2-Dichloropropane,
      Toluene,
      Ethyl benzene,  and
      Vinyl  chloride.

            7.3.5.1      If a %RSD greater than 30  percent  is measured for
      any  CCC,  then corrective action  to  eliminate  a  system leak and/or
      column reactive  sites is required before  reattempting  calibration.

      7.3.6  Linearity - If the %RSD of any compound is 15% or less, then
the  relative  response  factor  is   assumed  to   be   constant  over  the
calibration range, and the average relative  response  factor may be used
for quantitation.

            7.3.6.1      If the %RSD  of any compound is greater than 15%,
      construct  calibration   curves  of   area   ratio   (A/Ais)   versus
      concentration using first or higher order regression  fit of the five
      calibration  points.   The  analyst  should  select the regression order
      which  introduces the  least calibration error into the quantitation.
      The  use of calibration curves is a recommended alternative to average
      response factor  calibration  (Sec.  7.6.2.4),  and  a useful diagnostic
      of standard  preparation accuracy and  absorption activity  in  the
      chromatographic  system.

      7.3.7  These curves  are  verified each shift by purging a performance
standard.    Recalibration  is  required  only if calibration  and  on-going
performance criteria cannot be met.

7.4   GC/MS  calibration verification

      7.4.1  Prior to the analysis of samples,  inject  or purge 5-50 ng of
the 4-bromofluorobenzene standard following Method  5030.   The resultant
mass spectra for  the  BFB must meet all of the criteria given in Table 4
before sample  analysis begins.  These  criteria must be demonstrated each
12-hour shift.


                            8260A -  18                        Revision 1
                                                          September 1994

-------
      7.4.2 The initial calibration curve (Sec.  7,3)  for each compound of
interest must be checked and verified once  every 12 hours during analysis
with the introduction technique used  for samples.  This is accomplished by
analyzing  a calibration  standard  that is  at  a  concentration  near the
midpoint concentration for the working range of  the GC/MS by checking the
SPCC and CCC.

      7.4.3 System  Performance  Check  Compounds  (SPCCs)  -  A  system
performance check must  be made each  12 hours.   If the SPCC criteria are
met, a comparison of relative response factors is made  for all compounds.
This is the same check that is  applied during the initial calibration.  If
the minimum relative response  factors  are not  met,   the  system must be
evaluated,  and  corrective  action  must be taken  before  sample analysis
begins.    Some  possible  problems  are  standard  mixture  degradation,
injection port inlet contamination, contamination at  the front end of the
analytical  column,  and  active sites  in  the column  or  chromatographic
system.

            7.4.3.1      The minimum relative response factor for volatile
      SPCCs are  as  follows:

            Chloromethane                              0.10
            1,1-Dichloroethane                         0.10
            Bromoform                                 >0.10
            Chlorobenzene                              0.30
            1,1,2,2-Tetrachloroethane                  0.30

      7.4.4 Calibration  Check  Compounds  (CCCs)  -   After  the  system
performance check is met, CCCs listed in  Sec. 7.3.5 are used to check the
validity of the initial calibration.

      Calculate  the  percent drift  using the following equation:

                    % Drift =  (C, - CC)/C, x  100

where:

      C, =   Calibration Check Compound standard  concentration.
      Cc =   Measured concentration using selected quantitation  method.

      If the percent  drift for each  CCC  is less than 20%,  the initial
calibration is assumed to  be valid.   If  the criterion is  not met (> 20%
drift),  for any one  CCC,  corrective  action must be taken.   Problems
similar to  those listed  under  SPCCs  could  affect this criterion.   If no
source of the problem can be determined after corrective action has been
taken, a new  five  point calibration MUST  be generated.   This  criterion
MUST be met before quantitative sample analysis  begins.  If the CCCs are
not required analytes by the permit,  then all required analytes must meet
the 20% drift criterion.

      7.4.5 The internal  standard  responses and retention times  in the
check calibration standard must be evaluated immediately after or during
data acquisition.   If  the retention time for any  internal standard changes


                            8260A - 19                        Revision  1
                                                          September 1994

-------
by more than 30 seconds  from  the  last  calibration  check (12 hours),  the
chromatographic system must be inspected for malfunctions and corrections
must be made,  as required.   If the EICP  area  for any of  the internal
standards changes by a factor of two (-50% to +100%) from the last daily
calibration check standard, the mass spectrometer  must be inspected  for
malfunctions  and  corrections  must be  made,  as  appropriate.    When
corrections are made, reanalysis of samples analyzed while the system was
malfunctioning is necessary.

7.5  GC/MS  analysis

     7.5.1  It is highly  recommended that  the extract  be  screened on a
headspace-GC/FID  (Methods  3810/8015),  headspace-GC/PID/ELCD  (Methods
3810/8021),  or waste dilution-GC/PID/ELCD  (Methods  3585/8021)  using  the
same type of capillary column.  This will  minimize contamination of  the
GC/MS system from unexpectedly high concentrations of organic compounds.
Use of screening  is  particularly  important when this  method  is used to
achieve low detection levels.

     7.5.2  All samples and standard solutions must be  allowed to warm to
ambient temperature before analysis. Set up the purge-and-trap system as
outlined in Method 5030 if purge-and-trap introduction will be used.

     7.5.3  BFB  tuning  criteria   and   GC/MS  calibration  verification
criteria must be met before analyzing samples.

            7.5.3.1      Remove the  plunger from a  5 ml  syringe and attach
     a closed  syringe valve.  If lower  detection limits are required,- use
     a 25 ml syringe.  Open the sample  or standard bottle, which has been
     allowed to come to ambient temperature, and carefully  pour the sample
     into the  syringe  barrel  to just  short of overflowing.  Replace  the
     syringe plunger and compress the sample.  Open the syringe valve and
     vent any  residual  air while  adjusting the  sample  volume  to 5.0  ml.

     7.5.4  The process  of taking  an  aliquot  destroys the  validity of
aqueous and soil  samples for future analysis; therefore, if there is only
one VOA vial,  the analyst should prepare a second aliquot for analysis at
this time to  protect against possible  loss  of sample  integrity.   This
second sample  is  maintained only  until  such  time when the  analyst  has
determined that the first sample has been analyzed properly.  For aqueous
samples,  filling  one 20  ml  syringe would require  the use of  only  one
syringe.    If  a  second analysis  is needed  from  a  syringe, it  must be
analyzed within 24 hours.  Care  must be taken to prevent air from leaking
into the syringe.

            7.5.4.1      The   following   procedure  is  appropriate  for
     diluting  aqueous purgeable  samples.   All  steps  must  be performed
     without  delays  until  the diluted  sample  is in  a  gas-tight syringe.

                  7.5.4.1.1   Dilutions may be made in volumetric flasks
            (10 to 100 ml).  Select the volumetric flask that will allow
            for the  necessary dilution.   Intermediate dilutions  may be
            necessary for extremely large dilutions.


                            8260A  - 20                        Revision 1
                                                          September 1994

-------
                  7.5.4.1.2  Calculate the approximate volume of organic-
            free  reagent water  to  be   added  to  the  volumetric  flask
            selected and add slightly less  than this quantity of organic-
            free reagent water to the flask.

                  7.5.4.1.3  Inject  the proper aliquot of sample from the
            syringe into the flask.   Aliquots of less than 1 ml are not
            recommended.  Dilute the  sample to the mark with  organic-free
            reagent water.   Cap the flask,  invert, and shake  three times.
            Repeat above procedure for additional dilutions.

                  7.5.4.1.4  Fill a  5 ml syringe  with  the diluted sample.

            7.5.4.2      Compositing   aqueous   samples  prior  to  GC/MS
      analysis

                  7.5.4.2.1  Add  5  ml or  equal  larger  amounts  of each
            sample (up to 5  samples are allowed)  to a 25  mL glass syringe.
            Special  precautions must  be made to maintain  zero  headspace in
            the syringe.

                  7.5.4.2.2  The samples must  be  cooled at 4°C during this
            step to minimize volatilization losses.

                  7.5.4.2.3  Mix  well  and draw  out  a 5  ml  aliquot  for
            analysis.

                  7.5.4.2.4  Follow  sample  introduction,  purging,  and
            desorption steps described in Method  5030.

                  7.5.4.2.5  If  less than five samples  are  used  for
            compositing, a  proportionately smaller syringe  may  be used
            unless a 25 ml  sample is  to  be purged.

      7.5.5  Add  10.0  /xL  of surrogate  spiking   solution  and  10  /zL  of
internal standard  spiking  solution   to each  sample.   The  surrogate  and
internal standards may  be mixed  and  added  as  a  single spiking  solution.
The addition of 10 /nL  of the surrogate spiking solution  to 5  ml of sample
is equivalent to  a concentration  of  50 /ig/L  of  each  surrogate  standard.
The addition of 10 pi of the surrogate spiking solution to 5 g of sample
is equivalent to a concentration of  50 M9/kg of each surrogate standard.

            7.5.5.1       If  a more sensitive mass  spectrometer is employed
      to achieve lower detection levels, more dilute surrogate  and internal
      standard  solutions may be required.

      7.5.6  Perform purge-and-trap or direct injection  by Method 5030.  If
the  initial  analysis  of  sample or a   dilution of  the  sample  has  a
concentration of analytes that  exceeds the  initial calibration range,  the
sample  must  be  reanalyzed  at  a   higher  dilution.    Secondary  ion
quantitation is allowed only when there are sample interferences with the
primary ion.  When  a  sample is analyzed that has saturated  ions  from a
compound,  this analysis must be followed by a blank organic-free reagent


                            8260A -  21                        Revision 1
                                                          September 1994

-------
water analysis.  If the blank analysis is not free of interferences, the
system must be decontaminated.  Sample analysis may not resume until the
blank analysis is demonstrated to be free of interferences.

            7.5.6.1.    All  dilutions should  keep  the response  of the
      major constituents (previously saturated peaks) in the upper half of
      the  linear  range of the curve.  Proceed to Sees. 7.6.1 and 7.6.2 for
      qualitative and  quantitative  analysis.

      7.5.7 For  matrix  spike analysis,  add  10 juL  of  the matrix spike
solution (Sec. 5.13) to  the 5 ml of sample to be purged.  Disregarding any
dilutions, this is equivalent to a concentration of 50 p.g/1 of each matrix
spike standard.

7.6   Data interpretation

      7.6.1 Qualitative analysis

            7.6.1.1     The   qualitative  identification   of  compounds
      determined  by  this  method  is  based  on retention  time,   and  on
      comparison  of the sample mass  spectrum,  after background correction,
      with characteristic ions in a reference mass spectrum.  The reference
      mass  spectrum must  be  generated  by the  laboratory  using  the
      conditions  of  this method.    The  characteristic  ions  from  the
      reference mass spectrum are defined to be  the three ions of greatest
      relative intensity, or  any  ions over 30%  relative intensity if less
      than three such ions occur  in the reference  spectrum.   Compounds
      should be identified as  present  when  the  criteria below  are  met.

                   7.6.1.1.1  The intensities of  the characteristic ions
            of a compound maximize  in  the same  scan  or within one scan of
            each other.   Selection of  a peak by  a data  system  target
            compound search  routine where  the search  is  based  on  the
            presence of  a  target  chromatographic  peak containing ions
            specific  for  the  target  compound  at  a  compound-specific
            retention  time  will  be  accepted as  meeting  this criterion.

                   7.6.1.1.2  The RRT  of the sample component  is  within
            ± 0.06 RRT units  of the RRT of the  standard component.

                   7.6.1.1.3  The    relative    intensities    of    the
            characteristic   ions  agree  within  30%   of   the  relative
            intensities  of  these  ions  in  the   reference  spectrum.
            (Example:    For  an  ion with  an  abundance  of  50%  in  the
            reference  spectrum, the corresponding abundance  in  a  sample
            spectrum can range between 20% and  80%.)

                   7.6.1.1.4  Structural  isomers that produce very similar
            mass spectra should  be identified  as individual  isomers if
            they   have   sufficiently   different  GC   retention   times.
            Sufficient GC resolution  is achieved  if  the height  of the
            valley between two  isomer peaks is less than 25% of the  sum of
                            8260A - 22                        Revision 1
                                                          September 1994

-------
      the  two  peak  heights.   Otherwise,  structural  isomers  are
      identified  as  isomeric pairs.

             7.6.1.1.5  Identification  is  hampered  when   sample
      components  are not resolved chromatographically and  produce
      mass  spectra  containing  ions  contributed by  more than  one
      analyte.  When gas chromatographic  peaks  obviously represent
      more  than one  sample component (i.e., a  broadened  peak  with
      shoulder(s)  or a   valley  between   two   or   more maxima),
      appropriate  selection  of  analyte  spectra   and   background
      spectra  is  important.   Examination of extracted ion  current
      profiles  of appropriate  ions  can  aid  in the selection  of
      spectra,  and in qualitative  identification of compounds.  When
      analytes  coelute  (i.e.,  only  one chromatographic  peak  is
      apparent),  the identification  criteria  can be met, but  each
      analyte  spectrum will  contain  extraneous  ions  contributed by
      the coeluting  compound.

      7.6.1.2      For samples  containing components  not  associated
with the calibration  standards,  a  library search may be made  for the
purpose of tentative identification.   The necessity  to perform  this
type of  identification will  be  determined by  the type of analyses
being conducted. Guidelines for making tentative identification  are:

      (1)    Relative intensities of major ions  in  the  reference
             spectrum (ions > 10% of the most abundant ion)  should
             be present  in  the  sample spectrum.

      (2)    The relative intensities of the major ions should agree
             within ± 20%.   (Example:  For an ion with an  abundance
             of 50%  in  the  standard spectrum,  the corresponding
             sample ion  abundance  must be  between 30  and  70%).

      (3)    Molecular ions present in the reference  spectrum should
             be present  in  the  sample spectrum.

      (4)    Ions  present  in  the sample  spectrum  but not   in  the
             reference spectrum  should   be  reviewed  for possible
             background   contamination   or   presence  of   coeluting
             compounds.

      (5)    Ions  present  in the reference spectrum but not  in  the
             sample  spectrum  should  be  reviewed  for  possible
             subtraction   from   the   sample  spectrum  because   of
             background  contamination or coeluting  peaks.     Data
             system library reduction programs can sometimes  create
             these  discrepancies.

      Computer  generated  library search  routines should  not   use
normalization routines that would misrepresent the library or  unknown
spectra when compared to each other.   Only after visual  comparison
of sample with the nearest library searches will the mass spectral
interpretation specialist assign a tentative identification.
                      8260A - 23                        Revision 1
                                                    September 1994

-------
7.6.2 Quantitative  analysis

      7.6.2.1      When  a   compound   has  been   identified,   the
quantitation  of that  compound  will  be  based  on  the  integrated
abundance  from  the   EICP   of   the   primary   characteristic  ion.
Quantitation will take place using the internal standard technique.
The internal  standard  used  shall be the one nearest  the retention
time of that of a given analyte.

      7.6.2.2      When MS  response is linear and passes through the
origin,  calculate the concentration of each identified  analyte in the
sample as follows:

      Water

                                 (A,)(I.)
      concentration  (jug/L)  = 	zz;	
                              (AJ(RF)(V0)

where:

      Ax     =     Area of  characteristic  ion for compound  being
                   measured.
      Is     =     Amount of internal standard injected  (ng).
      Ais     =     Area  of  characteristic ion  for  the  internal
      	           standard.
      RF     =     Mean relative response factor for compound being
                   measured.
      V0     =     Volume  of   water   purged   (ml),   taking  into
                   consideration any dilutions made.

      Sediment/Soil  Sludge   (on  a  dry-weight  basis)  and  Waste
(normally on a wet-weight basis)

                                  (Ax)(U(Vt)
      concentration  (M9/kg)  = 	=r
                               (Ais)(RF)(Vi)(WJ(D)

where:

      Ax> Is>  Ais,  RF,  = Same as for water.
      Vt     =     Volume of total extract (/A) (use 10,000 juL or a
                   factor of this when dilutions are made).
      Vi     =     Volume of extract added (juL) for  purging.
      Ws     =     Weight of sample extracted or purged  (g).
      D     =     % dry weight of sample/100, or 1 for a wet-weight
                   basis.

      7.6.2.3      Where applicable, an estimate of concentration for
noncalibrated components in the sample  should be made.  The formulae
given above  should be  used  with  the  following modifications:  The
areas Ax and Ais should be from the total  ion  chromatograms, and the
RF for  the  compound  should be assumed to be 1.  The concentration

                       8260A -  24                         Revision 1
                                                     September 1994

-------
            obtained should be  reported indicating  (1)  that the value  is an
            estimate and  (2)  which  internal  standard  was  used  to  determine
            concentration.     Use  the  nearest   internal   standard   free  of
            interferences.

                  7.6.2.4     Alternatively, the regression  line fitted to the
            initial  calibration  (Sec.  7.3.6.1) may be used for determination of
            analyte  concentration.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and Method 8000 for  general  quality control
procedures.

      8.2   Additional  required  instrument QC is found in the  Sees. 7.3 and 7.4:

            8.2.1 The GC/MS system must be tuned to meet the BFB specifications.

            8.2.2 There must be  an initial  calibration of the GC/MS  system

            8.2.3 The  GC/MS  system  must meet  the SPCC  criteria and  the  CCC
      criteria, each 12 hours.

      8.3   To  establish   the  ability  to   generate  acceptable  accuracy  and
precision, the analyst must perform the following operations.

            8.3.1 A  quality control   (QC)  reference   sample  concentrate  is
      required containing each  analyte at a  concentration of  10 mg/L or less in
      methanol .  The QC reference sample concentrate may be prepared  from pure
      standard materials or purchased  as certified  solutions.   If prepared by
      the laboratory,  the  QC  reference sample concentrate must  be made  using
      stock standards prepared independently from those used for calibration.
            8.3.2  Prepare a QC reference sample to contain 20 ng/i or less of
      each analyte by adding 200 /zL of QC reference  sample concentrate to 100 mL
      of organic-free reagent water.

            8.3.3  Four  5-mL aliquots of the well  mixed QC reference sample are
      analyzed according to the method beginning  in Sec. 7.5.1.

            8.3.4  Calculate the average recovery  (x)  in  M9/L, and the standard
      deviation of the  recovery  (s)  in /zg/L, for  each  analyte  using the four
      results.

            8.3.5  Tables  7  and  8  provide  single  laboratory  recovery  and
      precision data obtained for  the method analytes  from water.   Similar
      results from dosed water should be expected by any experienced laboratory.
      Compare s and x   (Sec. 8.3.4) for  each  analyte  to the  single laboratory
      recovery and precision  data.   Results are comparable  if  the calculated
      standard deviation of the recovery does not  exceed 2.6  times the single
      laboratory RSD or 20%, whichever is greater, and  the mean recovery lies
      within the interval  x ± 3s  or x  ± 30%,  whichever is  greater.
                                  8260A - 25                        Revision 1
                                                                September 1994

-------
            NOTE:  The large  number of  analytes  in Tables  7  and 8  present  a
                  substantial probability that one or more  will  fail  at least
                  one of the acceptance criteria when all  analytes  of a given
                  method are determined.

            8.3.6  When one or more  of  the analytes tested are not comparable to
      the data  in Table 6  or  7,  the  analyst  must proceed according  to Sec.
      8.3.6.1 or 8.3.6.2.

                  8.3.6.1      Locate  and correct the source of the problem and
            repeat the test  for all  analytes  beginning with  Sec.  8.3.2.

                  8.3.6.2      Beginning with Sec.  8.3.2,  repeat the test only
            for  those  analytes that  are  not  comparable.   Repeated  failure,
            however,,  will confirm a general  problem with the measurement system.
            If  this occurs,  locate  and  correct the source  of  the  problem and
            repeat the test for all compounds  of interest  beginning with Sec.
            8.3.2.

      8.4   For  aqueous  and soil   matrices,  laboratory  established surrogate
control  limits should be compared with the control  limits  listed in Table 8.

            8.4.1  If recovery is not within limits,  the following procedures are
      required.

                  8.4.1.1      Check to  be sure that there are no errors in the
            calculations,  surrogate  solutions or internal standards.  If errors
            are  found,  recalculate  the data accordingly.

                  8.4.1.2      Check instrument performance.   If an instrument
            performance  problem is identified, correct the problem and re-analyze
            the  extract.

                  8.4.1.3      If no problem is found, re-extract and re-analyze
            the  sample.

                  8.4.1.4      If,  upon  re-analysis,  the recovery is again not
            within limits,  flag the data as "estimated concentration".

            8.4.2  At a minimum, each laboratory should update surrogate recovery
      limits on a matrix-by-matrix basis, annually.


9.0   METHOD PERFORMANCE

      9.1   The  method   detection   limit   (MDL)  is  defined  as  the  minimum
concentration of   a  substance   that  can  be  measured  and  reported with  99%
confidence that the value is above zero.  The MDL actually achieved in a given
analysis will vary depending on instrument sensitivity and matrix effects.

      9.2   This method  has  been  tested in  a single laboratory  using spiked
water.  Using a wide-bore capillary column, water was spiked at concentrations
between 0.5  and 10 M9/L.   Single  laboratory  accuracy and  precision  data are


                                  8260A - 26                        Revision 1
                                                                September 1994

-------
presented for the method analytes in Table  6.  Calculated MDLs are presented in
Table 1.

      9.3   The method was  tested  using water spiked  at  0.1 to 0.5  jug/L and
analyzed on a cryofocussed narrow-bore column.  The accuracy  and precision data
for these compounds are presented in Table 7-  MDL values were also calculated
from these data and are presented in Table 2.

      9.4   Direct injection has been used  for the analysis of waste motor oil
samples using  a  wide-bore column.   The accuracy  and  precision  data for these
compounds are presented in Table 10.


10.0  REFERENCES

1.    Methods for the  Determination  of Organic  Compounds  in Finished Drinking
      Water  and  Raw Source Water Method 524.2;  U.S.  Environmental  Protection
      Agency.  Office   of  Research  Development,  Environmental  Monitoring  and
      Support Laboratory, Cincinnati, OH 1986.

2.    U.S.  EPA Contract  Laboratory  Program,  Statement of  Work for Organic
      Analysis, July 1985, Revision.

3.    Bellar, T.A.;  J.J.  Lichtenberg. vh. Amer. Water Works Assoc. 1974, 66(12),
      739-744.

4.    Bellar,  T.A.;  J.J.  Lichtenberg.  "Semi-Automated Headspace Analysis  of
      Drinking Waters  and  Industrial  Waters for Purgeable Volatile Organic
      Compounds"; in Van Hall, Ed.; Measurement of Organic Pollutants in Water
      and Wastewater,  ASTM STP 686, pp 108-129,  1979.

5.    Budde, W.L.; J.W. Eichelberger.  "Performance Tests for the Evaluation of
      Computerized   Gas   Chromatography/Mass    Spectrometry  Equipment   and
      Laboratories";  U.S.  Environmental   Protection   Agency.    Environmental
      Monitoring and Support Laboratory, Cincinnati, OH 45268, April 1980; EPA-
      600/4-79-020.

6.    Eichelberger,  J.W.;  L.E.  Harris; W.L.  Budde.   "Reference Compound  to
      Calibrate   Ion   Abundance   Measurement   in    Gas   Chromatography-Mass
      Spectrometry Systems"; Analytical Chemistry 1975, 47, 995-1000.

7.    Olynyk, P.; W.L.  Budde;  J.W. Eichelberger. "Method  Detection Limit for
      Methods 624 and 625"; Unpublished report,  October 1980.

8.    Non   Cryogenic   Temperatures   Program   and   Chromatogram,   Private
      Communications;   Myron  Stephenson   and   Frank   Allen,  EPA   Region  IV
      Laboratory, Athens, GA.

9.    Marsden, P.; C.L. Helms, B.N. Colby.  "Analysis  of Volatiles  in  Waste Oil";
      report for B.  Lesnik, OSW/EPA under EPA contract 68-W9-001, 6/92.
                                  8260A - 27                        Revision 1
                                                                September 1994

-------
10.    Methods for  the  Determination of  Organic  Compounds in  Drinking  Water,
      Supplement II Method 524.2;  U.S.  Environmental  Protection Agency.  Office
      of Research and Development, Environmental Monitoring Systems Laboratory,
      Cincinnati, OH   1992.
                                  8260A - 28                        Revision 1
                                                                September 1994

-------
                            TABLE  1.
CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)
  FOR VOLATILE ORGANIC COMPOUNDS ON WIDE-BORE CAPILLARY COLUMNS
ANALYTE
Dichlorodifluoromethane
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Tri chl orof 1 uoromethane
Acrolein
lodomethane
Acetonitrile
Carbon disulfide
Allyl chloride
Methylene chloride
1,1-Dichloroethene
Acetone
trans-l,2-Dichloroethene
Acrylonitrile
1,1-Dichloroethane
Vinyl acetate
2,2-Dichloropropane
2-Butanone
cis-l,2-Dichloroethene
Propionitrile
Chloroform
Bromochloromethane
Methacrylonitrile
1,1, 1 -Tri chl oroethane
Carbon tetrachloride
1,1-Dichloropropene
Benzene
1,2-Dichloroethane
Trichloroethene
1,2-Dichloropropane
Bromodi chl oromethane
Dibromomethane
Methyl methacryl ate
1,4-Dioxane
2-Chloroethyl vinyl ether
4-Methyl -2-pentanone
trans-l,3-Dichloropropene
Toluene
cis-l,3-Dichloropropene
1,1,2-Tri chl oroethane
RETENTION TIME
(minutes)
Column la
1.35
1.49
1.56
2.19
2.21
2.42
3.19
3.56
4.11
4.11
4.11
4.40
4.57
4.57
4.57
5.00
6.14
6.43
8.10
--
8.25
8.51
9.01

9.19
10.18
11.02
--
11.50
12.09
14.03
14.51
15.39
15.43
15.50
16.17
--
17.32
17.47
18.29
19.38
19.59
Column 2"
0.70
0.73
0.79
0.96
1.02
1.19





2.06
1.57

2.36

2.93

3.80

3.90

4.80
4.38

4.84
5.26
5.29
5.67
5.83
7.27
7.66
8.49
7.93




--
10.00
--
11.05
Column 2'°
3.13
3.40
3.93
4.80

6.20





9.27
7.83

9.90

10.80

11.87

11.93

12.60
12.37

12.83
13.17
13.10
13.50
13.63
14.80
15.20
15.80
15.43




16.70
17.40
17.90
18.30
MDLd
(M9/L)
0.10
0.13
0.17
0.11
0.10
0.08





0.03
0.12

0.06

0.04

0.35

0.12

0.03
0.04

0.08
0.21
0.10
0.04
0.06
0.19
0.04
0.08
0.24




--
0.11
--
0.10
                           8260A - 29
    Revision 1
September 1994

-------
 TABLE 1.
(Continued)
ANALYTE
           RETENTION TIME
             (minutes)
         MDLd

Ethyl methacrylate
2-Hexanone
Tetrachl oroethene
1,3-Dichloropropane
Di bromochl oromethane
1,2-Dibromoethane
1-Chlorohexane
Chl orobenzene
1,1,1 , 2-Tetrachl oroethane
Ethyl benzene
p-Xylene
m-Xylene
o-Xylene
Styrene
Bromoform
Isopropyl benzene (Cumene)
cis-l,4-Dichloro-2-butene
1,1,2 , 2-Tetrachl oroethane
Bromobenzene
1,2,3-Trichloropropane
n-Propyl benzene
2-Chlorotoluene
trans-l,4-Dichloro-2-butene
1 , 3 , 5-Trimethyl benzene
4-Chlorotoluene
Pentachl oroethane
1,2, 4 -Tr i methyl benzene
sec-Butyl benzene
tert-Butyl benzene
p- Isopropyl tol uene
1 , 3-Di chl orobenzene
1,4-Dichlorobenzene
Benzyl chloride
n-Butyl benzene
1,2-Dichlorobenzene
l,2-Dibromo-3-chloropropane
1, 2, 4-Tri chl orobenzene
Hexachlorobutadiene
Naphthalene
1 , 2 , 3-Tri chl orobenzene
Column 1"
20.01
20.30
20.26
20.51
21.19
21.52
--
23.17
23.36
23.38
23.54
23.54
25.16
25.30
26.23
26.37
27.12
27.29
27.46
27.55
27.58
28.19
28.26
28.31
28.33
29.41
29.47
30.25
30.59
30.59
30.56
31.22
32.00
32.23
32.31
35.30
38.19
38.57
39.05
40.01
Column 2b


11.15
11.31
11.85
11.83
13.29
13.01
13.33
13.39
13.69
13.68
14.52
14.60
14.88
15.46

16.35
15.86
16.23
16.41
16.42

16.90
16.72

17.70
18.09
17.57
18.52
18.14
18.39

19.49
19.17
21.08
23.08
23.68
23.52
24.18
Column 2'°


18.60
18.70
19.20
19.40
--
20.67
20.87
21.00
21.30
21.37
22.27
22.40
22.77
23.30

24.07
24.00
24.13
24.33
24.53

24.83
24.77

31.50
26.13
26.60
26.50
26.37
26.60

27.32
27.43
--
31.50
32.07
32.20
32.97


0.14
0.04
0.05
0.06
0.05
0.04
0.05
0.06
0.13
0.05
0.11
0.04
0.12
0.15

0.04
0.03
0.32
0.04
0.04

0.05
0.06

0.13
0.13
0.14
0.12
0.12
0.03

0.11
0.03
0.26
0.04
0.11
0.04
0.03
 8260A -  30
    Revision 1
September 1994

-------
                                   TABLE 1.
                                  (Continued)
ANALYTE
RETENTION TIME
  (minutes)
                                     Column  la    Column 2    Column 2'c
 MDLd
(M9/L)
INTERNAL STANDARDS/SURROGATES
1,4-Difluorobenzene
Chlorobenzene-d5
l,4-Dichlorobenzene-d4
4-Bromofl uorobenzene
1 , 2-Di chl orobenzene-d4
Dichloroethane-d4
Di bromof 1 uoromethane
Toluene-d8
Pentaf 1 uorobenzene
Fl uorobenzene
13.26
23.10
31.16
27.83 15.71
32.30 19.08
12.08

18.27

13.00 6.27



23.63
27.25




14.06
a  Column 1 - 60 meter x 0.75 mm ID VOCOL capillary.   Hold  at 10°C for 8 minutes,
   then program  to  180°C at 4°/niin.

b  Column 2  - 30 meter x 0.53 mm  ID DB-624 wide-bore capillary using cryogenic
   oven.  Hold at 10°C for 5 minutes,  then program to 160°C at 6°/min.

c  Column 2' - 30 meter x 0.53 mm  ID DB-624 wide-bore capillary, cooling GC oven
   to ambient  temperatures.   Hold  at 10°C  for 6 minutes,  program  to  70°C  at
   10°/min,  program to 120°C  at  5°/min,  then  program  to  180°C  at 8°/min.

d  MDL based on  a 25 mL sample volume.
                                  8260A  - 31
                       Revision 1
                   September 1994

-------
                            TABLE  2.
CHROMATOGRAPHIC RETENTION TIMES AND METHOD DETECTION LIMITS (MDL)
 FOR VOLATILE ORGANIC COMPOUNDS ON NARROW-BORE CAPILLARY COLUMNS
ANALYTE
                              RETENTION TIME
                                (minutes)
                                Column 3a
MDLb
Di chl orodi f 1 uoromethane
Chl oromethane
Vinyl chloride
Bromomethane
Chloroethane
Tr i chl orof 1 uoromethane
1,1-Dichloroethene
Methylene chloride
trans- 1,2-Di chl oroethene
1,1-Dichloroethane
cis- 1,2-Di chl oroethene
2,2-Dichloropropane
Chloroform
Bromochl oromethane
1,1,1-Tri chl oroethane
1,2-Dichloroethane
1,1-Dichloropropene
Carbon tetrachloride
Benzene
1,2-Dichloropropane
Trichloroethene
Dibromomethane
Bromodi chl oromethane
Toluene
1,1, 2 -Tri chl oroethane
1,3-Dichloropropane
Di bromochl oromethane
Tetrachl oroethene
1,2-Di bromoethane
Chlorobenzene
1,1,1 , 2-Tetrachl oroethane
Ethyl benzene
p-Xylene
m-Xylene
Bromoform
o-Xylene
Styrene
1,1,2, 2-Tetrachl oroethane
1,2,3-Trichloropropane
Isopropyl benzene
0.88
0.97
1.04
1.29
1.45
1.77
2.33
2.66
3.54
4.03
5.07
5.31
5.55
5.63
6.76
7.00
7.16
7.41
7.41
8.94
9.02
9.09
9.34
11.51
11.99
12.48
12.80
13.20
13.60
14.33
14.73
14.73
15.30
15.30
15.70
15.78
15.78
15.78
16.26
16.42
0.11
0.05
0.04
0.06
0.02
0.07
0.05
0.09
0.03
0.03
0.06
0.08
0.04
0.09
0.04
0.02
0.12
0.02
0.03
0.02
0.02
0.01
0.03
0.08
0.08
0.08
0.07
0.05
0.10
0.03
0.07
0.03
0.06
0.03
0.20
0.06
0.27
0.20
0.09
0.10
                            8260A - 32
        Revision  1
   September  1994

-------
                                   TABLE 2.
                                  (Continued)
ANALYTE
RETENTION TIME
  (minutes)
  Column 3a
 MDLb
(M9/L)
Bromobenzene
2-Chlorotoluene
n-Propyl benzene
4-Chlorotoluene
1,3, 5-Trimethyl benzene
tert-Butyl benzene
1 , 2 , 4-Trimethyl benzene
sec-Butyl benzene
1,3-Dichlorobenzene
p-Isopropyl toluene
1 ,4-Dichlorobenzene
1,2-Dichlorobenzene
n-Butyl benzene
l,2-Dibromo-3-chloropropane
1,2,4-Trichlorobenzene
Naphthalene
Hexachlorobutadiene
1,2,3-Trichlorobenzene
16.42
16.74
16.82
16.82
16.99
17.31
17.31
17.47
17.47
17.63
17.63
17.79
17.95
18.03
18.84
19.07
19.24
19.24
0.11
0.08
0.10
0.06
0.06
0.33
0.09
0.12
0.05
0.26
0.04
0.05
0.10
0.50
0.20
0.10
0.10
0.14
a  Column 3-30 meter x 0.32 mm ID DB-5 capillary with 1 /im  film  thickness,

b  MDL based on a 25 ml sample volume.
                                  8260A - 33
                                Revision 1
                            September 1994

-------
                                 TABLE 3.
           ESTIMATED QUANTITATION LIMITS FOR VOLATILE ANALYTES8
                                     Estimated Quantitation Limits
                                       (All  Analytes in Table 1}
                                  Ground water         Low Soil/Sediment6
                                                            /zg/kg
    Purging 5 mL of water              5

    Purging 25 mL of water             1

    Soil/Sediment
   Estimated Quantitation Limit (EQL) - The lowest concentration that can be
   reliably  achieved within specified limits of precision and accuracy during
   routine   laboratory  operating  conditions. The  EQL  is  generally  5  to 10
   times the MDL.  However, it may  be nominally  chosen within these guidelines
   to simplify data  reporting. For many analytes  the EQL is selected from the
   lowest non-zero standard in the calibration  curve.  Sample EQLs are highly
   matrix-dependent.  The EQLs  listed herein  are provided for guidance and may
   not always be achievable.

   EQLs listed for soil/sediment are based  on  wet  weight.  Normally data are
   reported on a dry weight basis;  therefore,  EQLs will be higher, based on
   the percent dry weight in  each  sample.
             Other Matrices                      Factor0
             Water miscible liquid waste             50
             High-concentration soil and sludge     125
             Non-water miscible waste               500
°EQL  =   [EQL for low soil/sediment  (see Table 3)] X [Factor]. For non-aqueous
        samples, the factor is on a wet-weight basis.
                                8260A - 34                        Revision 1
                                                              September 1994

-------
                             TABLE 4.
    BFB MASS -  INTENSITY SPECIFICATIONS  (4-BROMOFLUOROBENZENE)'
Mass              Intensity Required (relative abundance)
 50               15 to 40% of mass 95
 75               30 to 60% of mass 95
 95               base peak, 100% relative abundance
 96               5 to 9% of mass 95
173               less than 2% of mass 174
174               greater than 50% of mass 95
175               5 to 9% of mass 174
176               greater than 95% but less than 101% of mass 174
177               5 to 9% of mass 176
     Alternate tuning criteria may  be  used (e.g. CLP, Method  524.2,  or
     manufacturers'  instructions), provided that method performance is not
     adversely affected.
                            8260A - 35                        Revision 1
                                                          September 1994

-------
                             TABLE 5.
    CHARACTERISTIC MASSES  (M/Z)  FOR PURGEABLE  ORGANIC  COMPOUNDS
Analyte
  Primary
Characteristic
    Ion
  Secondary
Characteristic
    Ion(s)
Acetone
Acetonitrile
Acrolein
Acrylonitrile
Allyl alcohol
Ally! chloride
Benzene
Benzyl chloride
Bromoacetone
Bromobenzene
Bromochl oromethane
Bromodichloromethane
Bromoform
Bromomethane
iso-Butanol
n-Butanol
2-Butanone
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon disulfide
Carbon tetrachloride
Chloral hydrate
Chloroacetonitrile
Chlorobenzene
1-Chlorobutane
Chi orodi bromomethane
Chloroethane
2-Chloroethanol
bis-(2-chloroethyl) sulfide
2-Chloroethyl vinyl ether
Chloroform
Chi oromethane
Chloroprene
3-Chloropropionitrile
2-Chlorotoluene
4-Chlorotoluene
1 , 2-Di bromo-3-chl oropropane
Di bromochl oromethane
1,2-Dibromoethane
Di bromomethane
1,2-Dichlorobenzene
l,2-Dichlorobenzene-d4
58
41
56
53
57
76
78
91
136
156
128
83
173
94
74
56
72
91
105
119
76
117
82
48
112
56
129
64(49*)
49
109
63
83
50(49*)
53
54
91
91
75
129
107
93
146
152
43
41, 40, 39
55, 58
52, 51
57, 58, 39
76, 41, 39, 78
-
91, 126, 65, 128
43, 136, 138, 93, 95
77, 158
49, 130
85, 127
175, 254
96
43
41
43, 72
92, 134
134
91, 134
78
119
44, 84, 86, 111
75
77, 114
49
208, 206
66(51*)
49, 44, 43, 51, 80
111, 158, 160
65, 106
85
52(51*)
53, 88, 90, 51
54, 49, 89, 91
126
126
155, 157
127
109, 188
95, 174
111, 148
115, 150
                            8260A - 36
                                Revision 1
                            September 1994

-------
                        TABLE 5.(continued)
Analyte
  Primary
Characteristic
    Ion
  Secondary
Characteristic
    Ion(s)
1,3-Dichlorobenzene
1 , 4-Di chl orobenzene
cis-l,4-Dichloro-2-butene
trans -1, 4-Di chl oro-2-butene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans -1,2-Di chl oroethene
1,2-Dichloropropane
1,3-Dichloropropane
2 , 2-Di chl oropropane
l,3-Dichloro-2-propanol
1,1-Dichloropropene
cis-l,3-Dichloropropene
trans- 1,3-Oi chl oropropene
1,2,3,4-Diepoxybutane
Diethyl ether
1,4-Dioxane
Epichlorohydrin
Ethanol
Ethyl acetate
Ethyl benzene
Ethylene oxide
Ethyl methacrylate
Hexachlorobutadiene
Hexachloroethane
2-Hexanone
2-Hydroxypropionitrile
lodomethane
Isobutyl alcohol
Isopropyl benzene
p- I sopropyl toluene
Malononitrile
Methacrylonitrile
Methyl acrylate
Methyl -t-butyl ether
Methylene chloride
Methyl ethyl ketone
Methyl iodide
Methyl methacrylate
4-Methyl -2-pentanone
Naphthalene
Nitrobenzene
146
146
75
53
85
63
62
96
96
96
63
76
77
79
75
75
75
55
74
88
57
31
88
91
44
69
225
201
43
44
142
43
105
119
66
41
55
73
84
72
142
69
100
128
123
HI,
111,
75,
88,
87
65,
98
61,
61,
61,
112
78
97
79,
110,
77,
77,
55,
45,
88,
57,
45,
43,
106
44,
69,
223,
166,
58,
44,
127,
43,
120
134,
66,
41,
85
57
86,
43
142,
69,
43,
-
51,
148
148
53,
75

83

63
98
98



43,
77
39
39
57,
59
58,
49,
27,
45,

43,
41,
227
199,
57,
43,
141
41,

91
39,
67,


49

127,
41,
58,

77


77, 124










81, 49



56

43, 57
62, 51
46
61

42
99, 86,

203
100
42, 53

42, 74


65, 38
39, 52,




141
100, 39
85




, 89






















114









66









                            8260A - 37
                                Revision 1
                            September 1994

-------
                        TABLE  5.(continued)
                                Primary           Secondary
                              Characteristic    Characteristic
Analyte                           Ion               Ion(s)
2-Nitropropane
2-Picoline
Pentachloroethane
Propargyl alcohol
6-Propiolactone
Propionitrile (ethyl cyanide)
n-Propylamine
n-Propyl benzene
Pyridine
Styrene
1,2,3-Trichlorobenzene
1,2,4-Trichlorobenzene
1,1,1 , 2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1,1,1-Trichl oroethane
1 , 1 , 2-Trichl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2,3-Trichloropropane
1 , 2 , 4-Trimethyl benzene
1, 3, 5-Trimethyl benzene
Vinyl acetate
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
INTERNAL STANDARDS/SURROGATES
1,4-Difluorobenzene
Chlorobenzene-d5
l,4-Dichlorobenzene-d4
4-Bromof 1 uorobenzene
Di bromof 1 uoromethane
Dichloroethane-d4
Toluene-d8
Pentafl uorobenzene
Fl uorobenzene
46
93
167
55
42
54
59
91
79
104
180
180
131
83
164
92
97
83
95
151
75
105
105
43
62
106
106
106

114
117
152
95
113
102
98
168
96

93,
167,
55,
42,
54,
59,
120
52
78
182,
182,
133,
131,
129,
91
99,
97,
97,
101,
77
120
120
86
64
91
91
91



115,
174,




77

66, 92, 78
130, 132, 165, 169
39, 38, 53
43, 44
52, 55, 40
41, 39



145
145
119
85
131, 166

61
85
130, 132
153











150
176





* - characteristic ion for an ion  trap mass spectrometer  (to be used when
ion-molecule reactions are observed)
                            8260A - 38                        Revision 1
                                                          September 1994

-------
                         TABLE 6.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR VOLATILE
    ORGANIC COMPOUNDS IN WATER DETERMINED WITH A WIDE-
                   BORE  CAPILLARY COLUMN
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodi chl oromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chl oromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-Chloropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1 , 2-Di chl orobenzene
1, 3 -Di chlorobenzene
1,4-Dichlorobenzene
Dichlorodifluoromethane
1 , 1 -Di chl orobenzene
1,2-Dichlorobenzene
1,1-Dichloroethene
cis-1, 2-Di chl oroethene
trans -1, 2-Di chl oroethene
1 , 2-Di chl oropropane
1,3-Dichloropropane
2 , 2-Di chl oropropane
1,1-Dichloropropene
Ethyl benzene
Hexachlorobutadiene
Isopropyl benzene
p-Isopropyl toluene
Methylene chloride
Naphthalene
n-Propyl benzene
Styrene
Cone. Number
Range, of Recovery8
jLtg/L Samples %
0.1
0.1
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.1
0.5
0.5
0.5
0.1
0.1
0.5
0.1
0.5
0.5
0.1
0.5
0.2
0.5
0.5
0.1
0.1
0.5
0.1
0.1
0.1
0.5
0.5
0.1
0.5
0.5
0.1
0.1
0.1
0.1
0.1
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 20
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
- 10
-100
- 10
-100
31
30
24
30
18
18
18
16
18
24
31
24
24
23
31
31
24
31
24
24
31
24
31
18
24
31
34
18
30
30
31
12
18
31
18
16
23
30
31
31
39
97
100
90
95
101
95
100
100
102
84
98
89
90
93
90
99
83
92
102
100
93
99
103
90
96
95
94
101
93
97
96
86
98
99
100
101
99
95
104
100
102
Standard
Deviation Percent
of Recovery13 RSD
6.5
5.5
5.7
5.7
6.4
7.8
7.6
7.6
7.4
7.4
5.8
8.0
5.5
8.3
5.6
8.2
16.6
6.5
4.0
5.6
5.8
6.8
6.6
6.9
5.1
5.1
6.3
6.7
5.2
5.9
5.7
14.6
8.7
8.4
6.8
7.7
6.7
5.0
8.6
5.8
7.3
5.7
5.5
6.4
6.1
6.3
8.2
7.6
7.6
7.3
8.8
5.9
9.0
6.1
8.9
6.2
8.3
19.9
7.0
3.9
5.6
6.2
6.9
6.4
7.7
5.3
5.4
6.7
6.7
5.6
6.1
6.0
16.9
8.9
8.6
6.8
7.6
6.7
5.3
8.2
5.8
7.2
                        8260A  -  39
    Revision 1
September 1994

-------
                                   TABLE 6.
                                  (Continued)

Analyte
Cone.
Range,
M9/L
Number
of Recovery"
Samples %
Standard
Deviation Percent
of Recovery1" RSD
1,1,1,2-Tetrachloroethane
1,1,2,2-Tetrachloroethane
Tetrachloroethene
Toluene
1,2,3-Tri chlorobenzene
1,2,4-Tri chlorobenzene
1,1,1-Trichloroethane
1,1,2-Trichloroethane
Trichloroethene
Tri chlorof1uoromethane
1,2,3-Trichloropropane
1,2,4-Trimethylbenzene
1,3,5-Trimethylbenzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene
0.5 -
0.1 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.5 -
0.1 -
0.1 -
0.5 -
10
10
10
10
10
10
10
10
10
10
10
10
10
10
31
10
10
24
30
24
18
18
18
18
18
24
24
16
18
23
18
18
31
18
 90
 91
 89
102
109
108
 98
104
 90
 89
108
 99
 92
 98
103
 97
104
  6.1
 5.7
 6.0
 8.1
 9.4
 9.0
 7.9
 7.6
 6.5
 7.2
15.6
 8.0
 6.8
 6.5
 7.4
 6.3
 8.0
 6.8
 6.3
 6.8
 8.0
 8.6
 8.3
 8.1
 7.3
 7.3
 8.1
14.4
 8.1
 7.4
 6.7
 7.2
 6.5
 7.7
a  Recoveries were calculated using internal  standard method.  Internal standard
   was fluorobenzene.

b  Standard deviation was calculated by pooling  data from three concentrations.
                                  8260A  - 40
                          Revision 1
                      September 1994

-------
                    TABLE 7.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR
 VOLATILE ORGANIC COMPOUNDS  IN WATER DETERMINED
       WITH A NARROW-BORE CAPILLARY COLUMN
Analyte
Benzene
Bromobenzene
Bromochl oromethane
Bromodichloromethane
Bromoform
Bromomethane
n-Butyl benzene
sec-Butyl benzene
tert-Butyl benzene
Carbon tetrachloride
Chlorobenzene
Chloroethane
Chloroform
Chi oromethane
2-Chlorotoluene
4-Chlorotoluene
l,2-Dibromo-3-chloropropane
Di bromochl oromethane
1,2-Dibromoethane
Dibromomethane
1, 2 -Di chlorobenzene
1, 3 -Di chlorobenzene
1, 4 -Di chlorobenzene
Di chl orodi f 1 uoromethane
1,1-Dichloroethane
1,2-Dichloroethane
1,1-Dichloroethene
cis-l,2-Dichloroethene
trans-l,2-Dichloroethene
1 ,2-Dichloropropane
1,3-Dichloropropane
2 , 2 -Di chl oropropane
1 , 1-Dichloropropene
Ethyl benzene
Hexachl orobutadi ene
Isopropyl benzene
p- I sopropyl to! uene
Methylene chloride
Naphthalene
n-Propyl benzene
Cone.
M9/L
0.1
0.5
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.1
0.1
0.5
0.5
0.5
0.5
0.1
0.5
0.5
0.1
0.1
0.1
0.1
0.5
0.1
0.1
0.1
0.1
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
0.5
Number
of
Samples
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
Recovery8
%
99
97
97
100
101
99
94
110
110
108
91
100
105
101
99
96
92
99
97
93
97
101
106
99
98
100
95
100
98
96
99
99
102
99
100
102
113
97
98
99
Standard
Deviation
of Recovery
6.2
7.4
5.8
4.6
5.4
7.1
6.0
7.1
2.5
6.8
5.8
5.8
3.2
4.7
4.6
7.0
10.0
5.6
5.6
5.6
3.5
6.0
6.5
8.8
6.2
6.3
9.0
3.7
7.2
6.0
5.8
4.9
7.4
5.2
6.7
6.4
13.0
13.0
7.2
6.6
Percent
RSD
6.3
7.6
6.0
4.6
5.3
7.2
6.4
6.5
2.3
6.3
6.4
5.8
3.0
4.7
4.6
7.3
10.9
5.7
5.8
6.0
3.6
5.9
6.1
8.9
6.3
6.3
9.5
3.7
7.3
6.3
5.9
4.9
7.3
5.3
6.7
6.3
11.5
13.4
7.3
6.7
                   8260A - 41
    Revision 1
September 1994

-------
TABLE 7.
(Continued)


Analyte
Styrene
1,1,1 , 2-Tetrachl oroethane
1,1,2 , 2-Tetrachl oroethane
Tetrachloroethene
Toluene
1 , 2 , 3-Tri chl orobenzene
1,2,4-Trichlorobenzene
1,1,1-Trichloroethane
1 , 1 , 2-Tri chl oroethane
Trichloroethene
Tri chl orof 1 uoromethane
1,2, 3-Tri chl oropropane
1,2, 4 -Tri methyl benzene
1 , 3 , 5-Tri methyl benzene
Vinyl chloride
o-Xylene
m-Xylene
p-Xylene

Cone.
M9/L
0.5
0.5
0.5
0.1
0.5
0.5
0.5
0.5
0.5
0.1
0.1
0.5
0.5
0.5
0.1
0.5
0.5
0.5
Number
of
Samples
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7

Recovery"
%
96
100
100
96
100
102
91
100
102
104
97
96
96
101
104
106
106
97
Standard
Deviation
of Recovery
19.0
4.7
12.0
5.0
5.9
8.9
16.0
4.0
4.9
2.0
4.6
6.5
6.5
4.2
0.2
7.5
4.6
6.1

Percent
RSD
19.8
4.7
12.0
5.2
5.9
8.7
17.6
4.0
4.8
1.9
4.7
6.8
6.8
4.2
0.2
7.1
4.3
6.3
Recoveries were calculated using internal  standard method. Internal standard
was fluorobenzene.
                               8260A - 42
    Revision 1
September 1994

-------
                                   TABLE 8.
      SURROGATE SPIKE RECOVERY LIMITS FOR WATER AND SOIL/SEDIMENT SAMPLES
Surrogate Compound
4-Bromof 1 uorobenzene"
Di bromof 1 uoromethane"
Toluene-d88
Dichloroethane-d4a
Percent
Low/High
Water
86-115
86-118
88-110
80-120
Recovery
Low/High
Soil/Sediment
74-121
80-120
81-117
80-120
8  Single laboratory data, for guidance only.
                                   TABLE 9.
                 QUANTITY OF EXTRACT REQUIRED FOR ANALYSIS OF
                          HIGH-CONCENTRATION SAMPLES
Approximate                                     Volume of
Concentration Range                             Extract3


   500 -  10,000 M9/kg                          100 /xL
 1,000 -  20,000 jug/kg                           50 ML
 5,000 - 100,000 M9/kg                           10 juL
25,000 - 500,000 /xg/kg                          100 /xL of  1/50 dilution6


Calculate appropriate dilution factor for concentrations exceeding this table.

a     The volume of solvent added to 5 mL of water being purged should be kept
      constant.  Therefore, add to the 5 mL syringe whatever volume of solvent
      is necessary to maintain a volume of 100 juL added to the syringe.

b     Dilute  an  aliquot  of the  solvent extract  and  then  take  100 juL  for
      analysis.
                                  8260A - 43                        Revision 1
                                                                September 1994

-------
                             TABLE 10
           DIRECT  INJECTION  ANALYSIS  OF  NEW OIL  AT  5
PPM

Compound
Acetone
Benzene
n-Butanol*,**
iso-Butanol*,**
Carbon tetrachloride
Carbon disulfide**
Chlorobenzene
Chloroform
1,4-Dichlorobenzene
1,2-Dichloroethane
1,1-Dichloroethene
Diethyl ether
Ethyl acetate
Ethyl benzene
Hexachloroethane
Methylene chloride
Methyl ethyl ketone
MIBK
Nitrobenzene
Pyridine
Tetrachloroethene

Recovery (%)
91
86
107
95
86
53
81
84
98
101
97
76
113
83
71
98
79
93
89
31
82
Trichlorofluoromethane 76
l,l,2-Cl3F3ethane
Toluene
Trichloroethene
Vinyl chloride
o-Xylene
m/p-Xylene
* Alternate mass
** TS nuant itatinr
69
73
66
63
83
84
employed
i

%RSD
14.8
21.3
27.8
19.5
44.7
22.3
29.3
29.3
24.9
23.1
45.3
24.3
27.4
30.1
30.3
45.3
24.6
31.4
30.3
35.9
27.1
27.6
29.2
21.9
28.0
35.2
29.5
29.5

Blank
(ppm)
1.9
0.1
0.5
0.9
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.2
0.0
0.0
0.4
0.0
0.0
0.0
0.0
0.0
0.0
0.6
0.0
0.0
0.4
0.6

Spike
(ppm)
5.0
0.5
5.0
5.0
0.5
5.0
5.0
6.0
7.5
0.5
0.7
5.0
5.0
5.0
3.0
5.0
5.0
5.0
2.0
5.0
0.7
5.0
5.0
5.0
0.5
0.2
5.0
10.0

Data are taken from Reference 9.
                            8260A - 44
          Revision  1
      September 1994

-------
                            FIGURE 1.
                         PURGING DEVICE
FOAMT1U*
                        BUT 1M IN 0-0.
                     •— 14MMO.O.
                        INLET 1*4 IN. 0.0.
EOT m IN. 0.0.
10 MM GLASS fftr
MCOWM POAQ8ITV
SAMPLE INLET

2-WAY SYNNQE VALVC

17 CM » GAUGE SVMNGE NEEDLE

6 MM O.O. BU88EH SEPTUM

INLET 1M IN. O.D.
                                                       1/16 IN O.O
                                                   /^STAINLESS STEEL
                                                      13X
                                                      MOLECULAH SIEVE
                                                      MJRQE GAS FILTER
                                                        PUMOEGAS
                                                        FVOWCXJNTIWL
                          8260A  - 45
                                    Revision  1
                               September 1994

-------
                            FIGURE 2.
TRAP  PACKING  AND CONSTRUCTION  TO  INCLUDE  DESORB  CAPABILITY
      PACKING DETAIL
CONSTRUCTION DETAIL
           r 5 MM OLABB WOOL
           77 CM SJOCA GEL
           1SCMTENAXQC
           *• i CM yn, ov-i
             9MMQLAMWOOL
             14 FT m»OOT
             ACSSTANCCtMME
             WNMPCDSOUO
                                                  OONTIO.ANO
                                                  FVK9MCTCM
                                                a 106 m. u>.
                                                ai
                                                STi
                         8260A - 46
                           Revision 1
                       September  1994

-------
                            FIGURE  3.
         SCHEMATIC OF PURGE-AND-TRAP DEVICE  - PURGE MODE
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
PURGE OAS
FLOW CONTROL
13X MOLECULAR
SIEVE FILTER
UOMO INJECTION PORTS

   — COLUMN OVEN
                                    UUUV-
                                               CONFIRMATORY COLUMN
                                              TO DETECTOR
                                               • ANALYTICAL COLUMN
                              OPTIONAL *PORT COLUMN
                              SELECTION VALVE
                               =	/- TRAP INLET
                                     TRAP
                                     22-C
                                PURGING
                                DEVICE
             NOTE
             ALL LINES BETWEEN TRAP
             AND QC SHOULD B€ HCATEO
             TO WC.
                            8260A  - 47
                            Revision  1
                        September  1994

-------
                           FIGURE 4.
        SCHEMATIC OF PURGE-AND-TRAP DEVICE - DESORB MODE
CARRIER GAS
FLOW CONTROL
PRESSURE
REGULATOR
UOUW INJECTION PORTS

    -COLUMN OVEN
                                    JUUV-
                                               CONRRMATORY COLUMN
                                              TO DETECTOR
                                               ANALYTICAL COLUMN
                             OPTIONAL *PORT COLUMN
                             SELECTION VALVE
                                      TRAP INLET
PURGE GAS
FLOW CONTROL
13X MOLECULAR
SIEVt FILTER
                                             NOTE:
                                             ALL UNES BETWEEN TRAP
                                             AND GC SHOULD BE HEATED
                                             TOBTC.
                           8260A - 48
                           Revision 1
                        September 1994

-------
  CO
  o
in o
o: u-
=3 o
  o

  CO
  <
  C5
               9.
                     3N3ZN3aOHOX3 d.  -C'3'
                   3N32N380MOTH3UU  -
               o.
               9
                       3N32N38CM01H3ia -CM
               O
               *.
               *
                   3N3SAJ.5
1   «
J   O
J   10
M   ••
a

5   ?

J   Z
0
®
CD
                        »
               O-
     UJ

     u
                    u   a
                                       3N3H130WOTM3WM131
                                         3NWi3WOWO"IH3 1 OOWOWf
                                       3NWH130M01H3IM1   t *I *T
                      3NWMX30MOTH3IO  -t *t

                3N3MX30MOTM3ia -»»t-*»

                              3N3"UHX3W

                                    -T'T
                                                     IAN i A
                                                                     ..o
                                                                     *
                                                 M
i
i
s
B
                                                                                     en
                                                                                   c cr>
                                                                                   o •-•
                                                                    > .0
                                                                    0) E
                                                                   a: a>
                                                                     *->
                                                                      o.
                                                                      0)
                                                                     CO
                                                                   o
                                                                   ID
                                                                   C\J
                                                                   00

-------
  00
  O
  QC
  O
10 O
=3 O
ca
  C3
  o
  CO
                                     3N3ZN38CMOTH3IMI
                                                 .T
                                                                                   O»
                                                                                 C Ol
                                                                                 O I-H
                                                                                 i- 0)
                                                                                 > JD
                                                                0)
                                                               CO
                            3N3ZN380«OTH3ia -3  X

                               -£••
                                                                     .9
                     3N3MAJ.8
                                           3NOH13WOM01H30WOM8I0
                                                     i
                                                     I
                                                     I
                                                     s
                                                     i
o
in
CM
00
                                          3NWH13WOWO-HOIaOWOMf
                    $
                                       3N3HX3QU01H3IM1
                    trt —  X O
                     • O    to.
S3   85
"5.   u «a
IK   oz
1     i
                                   3N3Mi30«OTM3IQ -2M-
                                     3N3H130W01H3XQ -Z'T-
                                            3QIM01H3 IAN I A

-------
   00
   0
   O

   UJ
cc u.
=> o
CJ3
   CJ3
   O
   n:
   o
   00
                                                                                                       •5 8
                                                                                                         S
                    (SI) $ zdfr'T  j^


;2=  >T30/3uadoad X T'T ^.
                                    >HW/w«dai
                                                                                              c  -Q
                                                                                              CD £
                                                                                              rf a)
                                                                                                +J
                                                                                                 Q.
                                                                                                 OJ
                                                                                                oo
                                                                                              
-------
: t- ac
I UJ UJ
          2
          0.
          in

          o
                SZ HH _

              -°?g
                                                                                                   en
                                                                                                 c en
                                                                                                 O i-l
                                                                                                •t- 0)
                                                                                                 > J3
                                                                                                 tt) E
                                                                                                o: w
                                                                                                   ••->
                                                                                                   a.
                                                                                                   Q)
                                                                                                   CO
                                                                                        if
  cc.


  X
CO,
=1
  s
                                                                          aa
                                                                         ••a-
                                                                                                CSJ
                                                                                                in
                                                                              CM
                                                                              CO

-------
                  FIGURE 9.
              LOW SOILS IMPINGER
PURGE INLET nrnNC
SAMPLE OUTLET UTTING
E                     "****„ ••^•^^^^^••^
                JSffi
   6mm 00 CLASS TUBING
                                    SEPTUM
                                       CAP
            40ml VIAL
                8260A - 53
    Revision 1
September 1994

-------
                                  METHOD 8260A
VOLATILE ORGANIC  COMPOUNDS  BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY  (GC/MS)
                           CAPILLARY  COLUMN  TECHNIQUE
                       (   Start   J

/ 7.1 \
/ Select N. Direct
S procedure N^ Injection ^
V^ for introducing J ^
>v^ sample into /
NGC/MS./
Purge-and-trap [
V


7.2 Set GC/MS
operating
conditions.
J,
7.3.1 Tune
GC/MS system
with BFB.
1
7.3.2 Assemble
purge-and-trap
device and prepare
calibration standards.
1
7.3.2.1 Perform
purge-and-trap
analysis.



k "

7.3.4 Calculate
RFs for
5 SPCCs.
,
,
7.3.5 Calculate
%RSD of RF
for CCCs.
J
1
7.4 Perform
calibration
verification.
^
,
7.5 Perform GC/MS
analysis utilizing
Methods 5030
or 8260.
^
r
7.6.1 Identify
analytes by
comparing the
sample and standard
mass spectra.
                                                  7.6.2 Calculate the
                                                   concentration of
                                                    each identified
                                                       analyte.
                                                    7.6.2.3 Report
                                                      all results.
                                                    f   Stop   J
                                   8260A - 54
    Revision 1
September 1994

-------
                                 METHOD 8270B

                       SEMIVOLATILE ORGANIC  COMPOUNDS BY
   GAS CHROMATOGRAPHY/MASS SPECTROHETRY (GC/MS): CAPILLARY COLUMN TECHNIQUE
1.0   SCOPE AND APPLICATION

      1.1   Method 8270 is used to determine the concentration of semivolatile
organic compounds in extracts prepared from all  types of solid waste matrices,
soils, and ground water.  Direct  injection  of a  sample  may be used in limited
applications.  The following compounds can be determined by this method:
Compounds
        Appropriate Preparation Techniques


CAS No8       3510
      3540/
3520  3541  3550  3580
Acenaphthene
Acenaphthene-d10 (I.S.)
Acenaphthylene
Acetophenone
2-Acetylaminofluorene
l-Acetyl-2-thiourea
Aldrin
2-Aminoanthraquinone
Aminoazobenzene
4-Aminobiphenyl
3-Amino-9-ethylcarbazole
Anilazine
Aniline
o-Anisidine
Anthracene
Aramite
Aroclor - 1016
Aroclor - 1221
Aroclor - 1232
Aroclor - 1242
Aroclor - 1248
Aroclor - 1254
Aroclor - 1260
Azinphos -methyl
Barban
Benzidine
Benzoic acid
Benz(a)anthracene
Benzo(b)fl uoranthene
Benzo ( k) f 1 uoranthene
Benzo (g , h , i ) peryl ene
Benzo(a)pyrene
83-32-9

208-96-8
98-86-2
53-96-3
591-08-2
309-00-2
117-79-3
60-09-3
92-67-1
132-32-1
101-05-3
62-53-3
90-04-0
120-12-7
140-57-8
12674-11-2
11104-28-2
11141-16-5
53469-21-9
12672-29-6
11097-69-1
11096-82-5
86-50-0
101-27-9
92-87-5
65-85-0
56-55-3
205-99-2
207-08-9
191-24-2
50-32-8
X
X
X
X
X
LR
X
X
X
X
X
X
X
X
X
HS(43)
X
X
X
X
X
X
X
HS(62)
LR
CP
X
X
X
X
X
X
X
X
X
NO
ND
ND
X
ND
ND
ND
X
ND
X
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
X
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
ND
X
X
X
X
X
X
X
X
ND
ND
ND
X
ND
ND
ND
ND
ND
X
ND
X
ND
X
X
X
X
X
X
X
ND
ND
CP
X
X
X
X
X
X
X
X
X
X
X
LR
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
CP
X
X
X
X
X
X
                                  8270B - 1
                                   Revision  2
                               September  1994

-------
Appropriate Preparation Techniques
Compounds
p-Benzoquinone
Benzyl alcohol
a-BHC
0-BHC
6-BHC
7-BHC (Lindane)
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Bromoxynil
Butyl benzyl phthalate
2-sec-Butyl-4,6-dinitrophenol
Captafol
Captan
Carbaryl
Carbofuran
Carbophenothion
Chlordane
Chlorfenvinphos
4-Chloroaniline
Chlorobenzilate
5-Chloro-2-methylanil ine
4-Chloro-3-methyl phenol
3-(Chloromethyl )pyridine
hydrochloride
1 -Chi oronaphthal ene
2-Chloronaphthalene
2-Chlorophenol
4-Chloro- 1,2 -phenyl enedi ami ne
4-Chloro-l ,3-phenylenediamine
4-Chlorophenyl phenyl ether
Chrysene
Chrysene-d12 (I.S.)
Coumaphos
p-Cresidine
Crotoxyphos
2-Cyclohexyl-4,6-dinitro-phenol
4, 4' -ODD
4,4'-DDE
4,4'-DDT
Demeton-0
Demeton-S
Diallate (cis or trans)
2,4-Diaminotoluene
CAS Noa
106-51-4
100-51-6
319-84-6
319-85-7
319-86-8
58-89-9
111-91-1
111-44-4
108-60-1
117-81-7
101-55-3
1689-84-5
85-68-7
88-85-7
2425-06-1
133-06-2
63-25-2
1563-66-2
786-19-6
57-74-9
470-90-6
106-47-8
510-15-6
95-79-4
59-50-7

6959-48-4
90-13-1
91-58-7
95-57-8
95-83-0
5131-60-2
7005-72-3
218-01-9

56-72-4
120-71-8
7700-17-6
131-89-5
72-54-8
72-55-9
50-29-3
298-03-3
126-75-0
2303-16-4
95-80-7
3510 3520
OE
X
X
X
X
X
X
X
X
X
X
X
X
X
HS(55)
HS(40)
X
X
X
X
X
X
X
X
X

X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
HS(68)
X
X
DC,OE(42)
ND
X
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
3540/
3541
ND
ND
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
ND
ND
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
3550
ND
X
X
X
X
X
X
X
X
X
X
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
X

ND
X
X
X
ND
ND
X
X
X
ND
ND
ND
ND
X
X
X
ND
ND
ND
ND
3580
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

X
X
X
X
ND
ND
X
X
X
X
X
X
LR
X
X
X
X
X
X
X
8270B - 2
    Revision 2
September 1994

-------
Compounds
        Appropriate Preparation Techniques


CAS No"      3510
      3540/
3520  3541 3550   3580
Dibenz(a,j)acridine
Di benz ( a , h) anthracene
Dibenzofuran
Dibenzo(a,e)pyrene
1 , 2-Di bromo-3-chl oropropane
Di-n-butyl phthalate
Dichlone
1 , 2-Di chl orobenzene
1 , 3 -Di chl orobenzene
1,4-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S)
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos
Dicrotophos
Dieldrin
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
Dihydrosaffrole
Dimethoate
3,3' -Dimethoxybenzidine
Dimethyl aminoazobenzene
7,12-Dimethylbenz(a)-
anthracene
3,3' -Dimethyl benzidine
a, a-Dimethyl phenethyl ami ne
2, 4-Dimethyl phenol
Dimethyl phthalate
1,2-Dinitrobenzene
1,3-Dinitrobenzene
1,4-Dinitrobenzene
4, 6-Dinitro- 2 -methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap
Dinoseb
Dioxathion
Diphenylamine
5,5-Diphenylhydantoin
1,2-Diphenylhydrazine
Di-n-octyl phthalate
Disulfoton
224-42-0
53-70-3
132-64-9
192-65-4
96-12-8
84-74-2
117-80-6
95-50-1
541-73-1
106-46-7

91-94-1
120-83-2
87-65-0
62-73-7
141-66-2
60-57-1
84-66-2
56-53-1
64-67-5
56312-13-1
60-51-5
119-90-4
60-11-7
57-97-6

119-93-7
122-09-8
105-67-9
131-11-3
528-29-0
99-65-0
100-25-4
534-52-1
51-28-5
121-14-2
606-20-2
39300-45-3
88-85-7
78-34-2
122-39-4
57-41-0
122-66-7
117-84-0
298-04-4
X
X
X
NO
X
X
OE
X
X
X
X
X
X
X
X
X
X
X
AW,OS(67)
LR
NO
HE,HS(31)
X
X
CP(45)

X
ND
X
X
X
X
HE(14)
X
X
X
X
CP,HS(28)
X
ND
X
X
X
X
X
ND
X
X
ND
X
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND

ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
ND
X
ND
X
X
ND
ND
X
ND
ND
ND
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
NO
ND
ND

ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
ND
X
ND
X
X
ND
ND
X
X
ND
ND
X
ND
X
X
X
X
X
X
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND

ND
ND
X
X
ND
ND
ND
X
X
X
X
ND
ND
ND
X
ND
X
X
ND
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
ND
X
LR
X
CP

X
X
X
X
X
X
X
X
X
X
X
CP
X
ND
X
X
X
X
X
                                  8270B - 3
                                   Revision  2
                               September  1994

-------
Compounds
        Appropriate Preparation Techniques


CAS No'      3510
      3540/
3520  3541 3550   3580
Endosulfan I
Endosulfan II
Endosulfan sulfate
Endrin
Endrin aldehyde
Endrin ketone
EPN
Ethion
Ethyl carbamate
Ethyl methanesulfonate
Ethyl parathion
Famphur
Fensulfothion
Fenthion
Fluchloral in
Fluoranthene
Fluorene
2-Fluorobiphenyl (surr.)
2-Fluorophenol (surr.)
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachl orobutadi ene
Hexachlorocyclopentadiene
Hexachl oroethane
Hexachl orophene
Hexachl oropropene
Hexamethyl phosphorarnide
Hydroquinone
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
Isosafrole
Kepone
Leptophos
Malathion
Maleic anhydride
Mestranol
Methapyrilene
Methoxychlor
3-Methyl chol anthrene
4,4'-Methylenebis
(2-chloroaniline)
4,4'-Methylenebis
(N,N-dimethyl ani line)
959-98-8
33213-65-9
1031-07-8
72-20-8
7421-93-4
53494-70-5
2104-64-5
563-12-2
51-79-6
62-50-0
56-38-2
52-85-7
115-90-2
55-38-9
33245-39-5
206-44-0
86-73-7
321-60-8
367-12-4
76-44-8
1024-57-3
118-74-1
87-68-3
77-47-4
67-72-1
70-30-4
1888-71-7
680-31-9
123-31-9
193-39-5
465-73-6
78-59-1
120-58-1
143-50-0
21609-90-5
121-75-5
108-31-6
72-33-3
91-80-5
72-43-5
56-49-5

101-14-4

101-61-1
X
X
X
X
X
X
X
X
DC(28)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
AW,CP(62)
X
X
NO
X
X
X
DC(46)
X
X
HS(5)
HE
X
X
X
X

OE,OS(0)

X
X
X
X
X
X
X
ND
ND
NO
ND
X
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND

ND

X
X
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND

ND

ND
X
X
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
ND
ND
ND
ND
X
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND

ND

ND
X
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X

LR

ND
                                   8270B -  4
                                   Revision 2
                               September 1994

-------
Compounds
        Appropriate Preparation Techniques


CAS Noa       3510
      3540/
3520  3541 3550   3580
Methyl methanesulfonate
2-Methylnaphthalene
2-Methyl-5-nitroaniline
Methyl parathion
2-Methyl phenol
3-Methyl phenol
4-Methyl phenol
2 -Methyl pyri dine
Mevinphos
Mexacarbate
Mi rex
Monocrotophos
Naled
Naphthalene
Naphthalene-d8 (I.S.)
1,4-Naphthoquinone
1-Naphthylamine
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine
Nitrobenzene
Nitrobenzene-d5 (surr.)
4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
5-Nitro-o-toluidine
Nitroquinoline-1 -oxide
N-Nitrosodi butyl ami ne
N-Nitrosodi ethyl ami ne
N-Nitrosodimethyl amine
N-Nitrosomethyl ethyl ami ne
N-Nitrosodi phenyl amine
N-Nitrosodi -n-propyl amine
N-Nitrosomorphol ine
N-Nitrosopi peri dine
N-Nitrosopyrrol idine
Octamethyl pyrophosphoramide
4,4'-Oxydianiline
Parathion
Pentachl orobenzene
66-27-3
91-57-6
99-55-8
298-00-0
95-48-7
108-39-4
106-44-5
109-06-8
7786-34-7
315-18-4
2385-85-5
6923-22-4
300-76-5
91-20-3

130-15-4
134-32-7
91-59-8
54-11-5
602-87-9
88-74-4
99-09-2
100-01-6
99-59-2
98-95-3

92-93-3
1836-75-5
88-75-5
100-02-7
99-55-8
56-57-5
924-16-3
55-18-5
62-75-9
10595-95-6
86-30-6
621-64-7
59-89-2
100-75-4
930-55-2
152-16-9
101-80-4
56-38-2
608-93-5
X
X
X
X
X
X
X
X
X
HE,HS(68)
X
HE
X
X
X
X
OS(44)
X
DE(67)
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
ND
X
X
LR
X
X
X
ND
X
X
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
X
X
X
ND
X
X
ND
ND
X
X
ND
ND
ND
ND
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
X
X
ND
ND
ND
ND
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
ND
ND
ND
ND
ND
X
X
X
ND
X
X
ND
ND
X
X
ND
ND
ND
ND
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
X
X
ND
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
LR
X
X
X
                                  8270B - 5
                                   Revision  2
                               September  1994

-------
Compounds
        Appropriate Preparation Techniques


CAS No"      3510
      3540/
3520  3541 3550   3580
Pentachloronitrobenzene
Pentachlorophenol
Perylene-d12 (I.S.)
Phenacetin
Phenanthrene
Phenanthrene-d10 (I.S.)
Phenobarbital
Phenol
Phenol -d6 (surr.)
1,4-Phenylenediamine
Phorate
Phosalone
Phosmet
Phosphamidon
Phthalic anhydride
2-Picoline
Piperonyl sulfoxide
Pronamide
Propylthiouracil
Pyrene
Pyridine
Resorcinol
Safrole
Strychnine
Sul fall ate
Terbufos
Terphenyl-d14(surr.)
1,2,4,5-Tetrachlorobenzene
2,3,4,6-Tetrachlorophenol
Tetrachlorvinphos
Tetraethyl dithiopyrophosphate
Tetraethyl pyrophosphate
Thionazine
Thiophenol (Benzenethiol )
Toluene diisocyanate
o-Toluidine
Toxaphene
2,4,6-Tribromophenol (surr.)
1 , 2 , 4-Trichl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
Trifluralin
2,4,5-Trimethylaniline
Trimethyl phosphate
82-68-8
87-86-5

62-44-2
85-01-8

50-06-6
108-95-2

106-50-3
298-02-2
2310-17-0
732-11-6
13171-21-6
85-44-9
109-06-8
120-62-7
23950-58-5
51-52-5
129-00-0
110-86-1
108-46-3
94-59-7
60-41-3
95-06-7
13071-79-9
1718-51-0
95-94-3
58-90-2
961-11-5
3689-24-5
107-49-3
297-97-2
108-98-5
584-84-9
95-53-4
8001-35-2

120-82-1
95-95-4
88-06-2
1582-09-8
137-17-7
512-56-1
X
X
X
X
X
X
X
DC(28)
DC(28)
X
X
HS(65)
HS(15)
HE(63)
CP,HE(1)
ND
X
X
LR
X
ND
DC,OE(10)
X
AW,OS(55)
X
X
X
X
X
X
X
X
X
X
HE(6)
X
X
X
X
X
X
X
X
HE(60)
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
X
ND
ND
ND
ND
ND
X
X
X
X
X
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
ND
X
ND
ND
ND
ND
X
X
ND
X
X
ND
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
X
X
X
X
ND
ND
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
X
CP
ND
X
X
LR
X
ND
X
X
X
X
X
X
X
X
X
ND
X
X
X
X
X
X
X
X
X
X
X
X
X
                                   8270B -  6
                                   Revision 2
                               September 1994

-------
                                         Appropriate Preparation Techniques

                                                             3540/
Compounds                        CAS No8      3510     3520   3541 3550  3580
1,3,5-Trinitrobenzene 99-35-4
Tris(2,3-dibromopropyl) phosphate 126-72-7
Tri-p-tolyl phosphate 78-32-0
0,0,0-Triethyl phosphorothioate 126-68-1
X
X
X
X
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
X
LR
X
X
a     Chemical Abstract Service Registry Number.

AW =  Adsorption to walls of glassware during extraction and storage.
CP =  Nonreproducible chromatographic performance.
DC =  Unfavorable  distribution  coefficient (number in  parenthesis  is percent
      recovery).
HE =  Hydrolysis during  extraction accelerated by acidic  or  basic  conditions
      (number in parenthesis is percent recovery).
HS =  Hydrolysis during storage (number in parenthesis  is percent stability).
LR =  Low response.
ND =  Not determined.
OE =  Oxidation during  extraction accelerated by basic  conditions  (number in
      parenthesis  is percent recovery).
OS =  Oxidation during storage (number in parenthesis is percent stability).
X =   Greater than 70 percent recovery by this technique.


      1.2   Method  8270  can be used  to  quantitate most neutral,  acidic,  and
basic organic compounds that are  soluble  in methylene  chloride and  capable of
being eluted without derivatization as sharp  peaks  from a  gas chromatographic
fused-silica capillary  column  coated  with  a slightly  polar   silicone.   Such
compounds include polynuclear aromatic hydrocarbons, chlorinated hydrocarbons and
pesticides,  phthalate esters, organophosphate  esters, nitrosamines, haloethers,
aldehydes,  ethers,  ketones, anilines, pyridines,  quinolines,  aromatic  nitro
compounds,  and  phenols,  including nitrophenols.   See  Table  1 for a  list of
compounds and their characteristic ions that have been evaluated on the specified
GC/MS system.

      1.3   The following  compounds  may  require special treatment  when  being
determined by  this  method.   Benzidine can be subject to oxidative losses during
solvent  concentration.   Also,  chromatography  is  poor.   Under the  alkaline
conditions of the extraction step, a-BHC,  7-BHC, Endosulfan  I and II, and Endrin
are subject to decomposition.   Neutral  extraction should be performed if these
compounds  are  expected.    Hexachlorocyclopentadiene  is  subject  to  thermal
decomposition  in the inlet of the gas chromatograph,  chemical reaction in acetone
solution, and  photochemical  decomposition.  N-nitrosodimethylamine is difficult
to separate from the  solvent  under the chromatographic conditions  described.
N-nitrosodiphenylamine decomposes  in the gas chromatographic inlet and cannot be
separated   from   diphenylamine.       Pentachlorophenol,   2,4-dinitrophenol,


                                   8270B -  7                         Revision 2
                                                                September 1994

-------
4-nitrophenol, 4,6~dinitro-2-methylphenol,4-chloro-3-methylphenol, benzoicacid,
2-nitroaniline,  3-nitroaniline,  4-chloroaniline,  and benzyl alcohol are subject
to erratic chromatographic  behavior, especially if the  GC system is contaminated
with high boiling material.

      1.4    The   estimated   quantitation  limit   (EQL)   of  Method  8270  for
determining  an  individual  compound  is approximately 1 mg/kg  (wet  weight)  for
soil/sediment samples,  1-200 mg/kg for wastes (dependent on  matrix and method of
preparation), and 10 jug/L for ground water samples  (see Table 2).  EQLs will be
proportionately  higher for  sample  extracts  that  require  dilution to  avoid
saturation of the detector.

      1.5    This  method  is restricted to  use  by or under the  supervision of
analysts experienced  in  the  use  of gas  chromatograph/mass  spectrometers  and
skilled in the interpretation of mass spectra.  Each analyst  must demonstrate the
ability to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1    Prior  to using  this  method, the  samples  should be  prepared  for
chromatography using the  appropriate sample preparation and  cleanup  methods.
This  method describes  chromatographic  conditions that  will   allow  for  the
separation  of  the  compounds in  the extract  and  for  their qualitative  and
quantitative analysis by mass spectrometry.


3.0   INTERFERENCES

      3.1    Raw  GC/MS  data  from  all  blanks,   samples,  and spikes  must  be
evaluated for interferences.  Determine  if the  source of interference is in the
preparation and/or cleanup of the samples  and take corrective action to eliminate
the problem.

      3.2    Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.   To reduce  carryover,  the
sample syringe must  be rinsed out between samples with solvent.   Whenever an
unusually  concentrated  sample is encountered,  it should  be followed by  the
analysis of solvent to check for cross contamination.


4.0   APPARATUS AND MATERIALS

      4.1    Gas  chromatograph/mass  spectrometer  system

             4.1.1  Gas  chromatograph  -  An analytical   system complete with  a
      temperature-programmable   gas  chromatograph  suitable   for  splitless
      injection  and  all  required accessories,  including  syringes,  analytical
      columns, and gases.   The  capillary column  should be  directly coupled to
      the  source.
                                   8270B  -  8                         Revision 2
                                                                September 1994

-------
            4.1.2  Column  - 30 m x 0.25 mm ID (or 0.32 mm ID)  1 /im film thickness
      silicone-coated  fused-silica  capillary column  (J&W  Scientific  DB-5 or
      equivalent).

            4.1.3  Mass  spectrometer -  Capable of scanning  from  35  to 500 amu
      every 1  sec or  less,  using 70 volts  (nominal)  electron energy  in the
      electron  impact  ionization  mode.  The  mass  spectrometer must  be capable
      of  producing a mass  spectrum for decafluorotriphenylphosphine (DFTPP)
      which meets all of the criteria in Table 3 when 1 jttL of the GC/MS tuning
      standard  is  injected through the GC (50 ng of DFTPP).

            4.1.4  GC/MS interface -  Any GC-to-MS interface that gives acceptable
      calibration points  at  50  ng per  injection  for each compound of interest
      and achieves acceptable tuning performance  criteria may be  used.   For a
      narrow-bore  capillary  column,  the  interface is usually capillary-direct
      into the mass spectrometer  source.

            4.1.5  Data system - A computer system must be interfaced  to the mass
      spectrometer.   The  system must  allow the  continuous acquisition  and
      storage on machine-readable media  of all mass  spectra obtained throughout
      the duration  of the  chromatographic  program.   The  computer  must  have
      software that can search any GC/MS data file for ions of a specific mass
      and that  can plot such  ion  abundances  versus  time  or scan  number.   This
      type  of  plot  is  defined  as  an  Extracted  Ion Current Profile (EICP).
      Software must also be available that allows  integrating the abundances in
      any EICP  between specified  time or scan-number limits.   The most recent
      version of the EPA/NIST Mass Spectral  Library should  also be available.

            4.1.6  Guard column  (optional)  (J&W  Deactivated Fused Silica,  0.25
      mm ID x 6 m,  or equivalent)  between the  injecti'on port and the analytical
      column joined  with  column  joiners  (Hewlett  Packard No.   5062-3556,  or
      equivalent).

      4.2   Syringe  -  10  ml.

      4.3   Volumetric flasks,  Class A - Appropriate  sizes  with  ground  glass
stoppers.

      4.4   Balance  - Analytical, 0.0001 g.

      4.5   Bottles  - glass with  Teflon-lined screw caps  or crimp tops.


5.0   REAGENTS

      5.1   Reagent  grade inorganic chemicals  shall  be  used  in  all  tests.
Unless otherwise indicated, it is intended that all reagents shall conform to the
specifications  of the Committee  on Analytical  Reagents of  the American Chemical
Society, where  such  specifications  are  available.  Other  grades  may be  used,
provided it  is  first ascertained that the reagent  is  of sufficiently high purity
to permit its  use without lessening  the accuracy of the determination.
                                   8270B  - 9                         Revision 2
                                                                September 1994

-------
      5.2   Organic-free reagent water - All references to water  in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Stock  standard solutions (1000 mg/L)  -  Standard  solutions can be
prepared from pure standard materials or purchased as certified solutions.

            5.3.1  Prepare  stock standard solutions by accurately weighing about
      0.0100 g  of  pure material.   Dissolve the  material  in pesticide quality
      acetone  or  other  suitable  solvent  and  dilute to  volume in  a  10  ml
      volumetric flask.  Larger volumes  can be  used  at the convenience of the
      analyst.  When compound purity  is assayed to be  96% or greater, the weight
      may be used without correction  to calculate the  concentration of the stock
      standard.    Commercially  prepared  stock  standards may  be used  at  any
      concentration  if  they  are  certified  by  the  manufacturer  or by  an
      independent  source.

            5.3.2  Transfer the stock standard solutions into bottles with Teflon
      lined screw-caps.  Store at -10°C to -20°C or less and protect from light.
      Stock  standard  solutions  should  be checked  frequently  for signs  of
      degradation or evaporation, especially just prior to preparing calibration
      standards from them.

            5.3.3  Stock  standard  solutions must  be  replaced after 1  year or
      sooner  if comparison  with  quality  control check samples indicates  a
      problem.

      5.4   Internal standard solutions  - The internal standards recommended are
l,4-dichlorobenzene-d4,   naphthalene-d8,   acenaphthene-d10,   phenanthrene-d10,
chrysene-d12,  and perylene-d12  (see  Table 5).   Other  compounds may  be  used as
internal standards  as  long as the  requirements  given in  Sec. 7.3.2  are met.
Dissolve 0.200  g of each  compound  with a  small  volume  of carbon  disulfide.
Transfer to a 50 ml volumetric flask and dilute to volume with methylene chloride
so that the final  solvent  is  approximately  20%  carbon disulfide.   Most of the
compounds are also soluble in  small  volumes of  methanol, acetone, or toluene,
except for perylene-d12.   The resulting  solution will contain each standard at
a concentration of 4,000 ng//xL.  Each  1  ml sample extract undergoing analysis
should be spiked with 10 /iL  of the  internal  standard solution,  resulting in a
concentration of 40 ng/juL  of  each  internal  standard.   Store at -10°C  to -20°C
or less when not being used.

      5.5   GC/MS  tuning  standard  - A methylene  chloride solution  containing
50 ng//iL  of decafluorotriphenylphosphine  (DFTPP) should  be  prepared.    The
standard should also contain 50 ng//il_ each of 4,4'-DDT, pentachlorophenol, and
benzidine to verify injection port inertness and GC column performance.  Store
at -10°C to -20°C or less when not being used.

      5.6   Calibration  standards  - A  minimum  of five  calibration  standards
should  be  prepared.    One  of  the  calibration  standards  should  be  at  a
concentration near,  but  above,  the method detection  limit;  the  others should
correspond to the range of concentrations found in real samples but should not
exceed the working  range of the GC/MS system.  Each standard should contain each
analyte for detection by this method (e.g. some or all of the compounds listed
in Table 1 may be  included).   Each 1 ml aliquot of calibration standard should


                                  8270B - 10                        Revision 2
                                                                September 1994

-------
be spiked with 10 jxL of the internal standard solution prior to analysis.  All
standards should  be  stored at  -10°C to  -20°C  or  less,  and should  be  freshly
prepared once a year, or sooner if check standards indicate a problem.  The daily
calibration standard should be prepared weekly and stored at 4°C.

      5.7   Surrogate  standards  -  The  recommended  surrogate  standards  are
phenol-d6,    2-fluorophenol,    2,4,6-tribromophenol,    nitrobenzene-d5,
2-fluorobiphenyl, and p-terphenyl-du.  See Method 3500 for the instructions on
preparing the surrogate standards. Determine what concentration should be  in the
blank extracts after all  extraction, cleanup,  and concentration steps.   Inject
this concentration into the GC/MS to determine recovery of surrogate standards
in all  blanks, spikes,  and sample extracts.  Take into account all  dilutions of
sample extracts.

      5.8   Matrix  spike  standards -  See Method  3500  for  instructions  on
preparing the matrix  spike standard. Determine what concentration should be in
the  blank  extracts after  all  extraction, cleanup,  and  concentration  steps.
Inject this  concentration  into the GC/MS  to  determine recovery  of  surrogate
standards in  all  matrix  spikes.   Take into  account  all  dilutions  of  sample
extracts.

      5.9   Acetone, hexane, methylene chloride,  isooctane,  carbon  disulfide,
toluene,  and other appropriate solvents -  Pesticide quality or equivalent


6.0   SAMPLE COLLECTION,  PRESERVATION,  AND HANDLING

      6.1   See  the  introductory material  to this  chapter,  Organic  Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Sample  preparation  -  Samples  must  be  prepared  by  one  of  the
following methods prior to GC/MS analysis.

      Matrix                              Methods
      Water                               3510, 3520
      Soil/sediment                       3540, 3541, 3550
      Waste                               3540, 3541, 3550, 3580

            7.1.1 Direct  injection  -  In   very  limited  applications  direct
      injection of the sample into  the GC/MS system with a 10 /^L syringe may be
      appropriate.     The  detection   limit   is   very  high   (approximately
      10,000 jug/L);  therefore,  it  is only  permitted where  concentrations  in
      excess of  10,000  /ig/L are expected.   The  system must be  calibrated  by
      direct injection.
                                  8270B - 11                         Revision  2
                                                                September 1994

-------
      7.2   Extract cleanup - Extracts may be cleaned up by any of the following
methods prior to GC/MS analysis.

      Compounds                           Methods
      Phenols                             3630, 3640, 8040a
      Phthalate esters                    3610, 3620, 3640
      Nitrosamines                        3610, 3620, 3640
      Organochlorine pesticides & PCBs    3620, 3660
      Nitroaromatics and cyclic ketones   3620, 3640
      Polynuclear aromatic hydrocarbons   3611, 3630, 3640
      Haloethers                          3620, 3640
      Chlorinated hydrocarbons            3620, 3640
      Organophosphorus pesticides         3620
      Petroleum waste                     3611, 3650
      All priority pollutant base,
          neutral, and acids              3640

      8     Method 8040 includes a derivatization technique followed by GC/ECD
            analysis,   if interferences are encountered on GC/FID.

      7.3   Initial calibration - The  recommended GC/MS operating conditions:

      Mass range:             35-500 amu
      Scan time:              1 sec/scan
      Initial temperature:    40°C,  hold for 4  minutes
      Temperature program:    40-270°C at 10°C/min
      Final temperature:      270°C,  hold until benzo[g,h,i]perylene has eluted
      Injector temperature:   250-300°C
      Transfer line temperature: 250-300°C
      Source temperature:     According to manufacturer's specifications
      Injector:               Grob-type, splitless
      Sample volume:          1-2 /jL
      Carrier gas:            Hydrogen at 50 cm/sec or helium at 30 cm/sec

      (Split injection is  allowed if the sensitivity of the mass spectrometer
      is sufficient).

            7.3.1  Each GC/MS system must be  hardware-tuned to meet the criteria
      in Table  3  for a 50 ng  injection of  DFTPP.  Analyses should  not begin
      until  all  these  criteria are  met.    Background  subtraction  should  be
      straightforward and designed only to eliminate  column  bleed or instrument
      background ions.  The GC/MS  tuning standard  should also be used to assess
      GC column performance  and injection port inertness.  Degradation of DDT
      to DDE and ODD should not exceed 20%.   (See  Sec. 8.3.1 of Method 8081 for
      the  percent breakdown  calculation).    Benzidine and pentachlorophenol
      should be present at their  normal  responses,  and  no peak tailing should
      be visible.   If degradation is  excessive and/or  poor chromatography is
      noted, the injection port may  require  cleaning.  It  may also be necessary
      to break  off  the  first 6-12 in. of the  capillary column.   The use of a
      guard column  (Sec.  4.1.6) between the injection  port and the analytical
      column may help prolong analytical column performance.
                                  8270B - 12                        Revision 2
                                                                September 1994

-------
       7.3.2  The  internal standards selected in Sec. 5.4 should permit most
of the components  of interest  in a chromatogram to have retention times
of 0.80-1.20 relative to one of the internal standards.  Use the  base peak
ion  from  the   specific  internal  standard  as  the  primary  ion  for
quantitation (see Table 1).   If interferences are noted, use the  next most
intense ion as the quantitation ion (i.e.  for  l,4-dichlorobenzene-d4, use
152 m/z for quantitation).

       7.3.3  Analyze  1 /j,L of each calibration standard (containing internal
standards) and tabulate the  area of the primary characteristic ion against
concentration for each compound (as indicated in Table 1).  Figure 1 shows
a chromatogram of a calibration standard containing  base/neutral and acid
analytes.  Calculate response factors  (RFs) for  each compound relative to
one of the internal  standards as follows:

                         RF  - (AxCis)/(AisCx)

where:

       Ax     =     Area of the characteristic  ion  for the compound being
                  measured.
       Ais     =     Area of the characteristic ion  for the specific internal
                  standard.
       Cis     =     Concentration of the specific internal standard (ng//iL).
       Cx     =     Concentration of the compound  being measured (ng//ul_).

       7.3.4  A  system performance check must be  performed to ensure that
minimum average  RFs  are met before the calibration  curve  is used.   For
semivolatiles,  the  System  Performance   Check  Compounds   (SPCCs)  are:
N-nitroso-di-n-propylamine;hexachlorocyclopentadiene;2,4-dinitro-phenol;
and 4-nitrophenol.  The minimum acceptable  average RF  for these compounds
is 0.050.  These SPCCs typically have very low RFs  (0.1-0.2) and tend to
decrease in response as the  chromatographic system begins to deteriorate
or the  standard  material begins to deteriorate.    They  are usually the
first  to show  poor performance.   Therefore, they must meet the minimum
requirement when the system  is calibrated.

             7.3.4.1     The  percent  relative  standard  deviation  (%RSD)
       should be  less than  15%  for  each compound.   However, the %RSD for
       each individual Calibration Check Compound (CCC) (see Table 4) must
       be less than 30%.  The relative  retention  times  of  each compound in
       each calibration run   should agree within  0.06 relative retention
       time  units.    Late-eluting  compounds  usually  have  much  better
       agreement.

               SD
      %RSD =—— x 100
               RF
where:

       RSD    =     relative  standard deviation.
       RF     =     mean of 5  initial RFs for a  compound.
      SD     =     standard  deviation of average  RFs  for a compound.


                            8270B - 13                        Revision 2
                                                          September 1994

-------
                            - RF)'
      SD =
                    1=1  N - 1
            where:
                  RFj    = RF for each of the 5 calibration levels
                  N     = Number of RF values (i.e.,  5)

            7.3.4.2     If the %RSD of any CCC is 30% or greater, then the
      chromatographic system is  too reactive for analysis  to begin.  Clean
      or replace the injector liner and/or capillary column,  then repeat
      the calibration procedure beginning with section 7.3.

      7.3.5 Linearity -  If the %RSD of any compound is 15% or less, then
the  relative  response   factor   is   assumed  to  be  constant  over  the
calibration range, and the average relative  response  factor  may be used
for quantitation (Sec. 7.6.2).

            7.3.5.1     If the  %RSD of any compound is greater than 15%,
      construct   calibration   curves  of   area  ratio   (A/Ais)   versus
      concentration using first  or higher order regression fit of the five
      calibration points.  The  analyst should select the regression order
      which introduces the least calibration error into the quantitation
      (Sec. 7.6.2.2  and  7.6.2.3).    The  use  of calibration curves  is  a
      recommended alternative to average  response factor calibration, and
      a useful diagnostic of  standard preparation accuracy and absorption
      activity in the chromatographic system.

7.4   Daily GC/MS calibration

      7.4.1 Prior to analysis of samples, the GC/MS tuning standard must
be analyzed.  A 50 ng injection  of DFTPP must  result  in  a mass spectrum
for DFTPP which meets the criteria given  in Table 3.  These criteria must
be demonstrated during each 12  hour shift.

      7.4.2 A calibration standard(s) at  mid-concentration containing all
semivolatile  analytes,   including  all   required  surrogates,   must  be
analyzed every 12  hours during analysis.  Compare the instrument response
factor from the standards  every  12 hours with  the  SPCC  (Sec.  7.4.3) and
CCC (Sec. 7.4.4) criteria.

      7.4.3 System   Performance  Check   Compounds  (SPCCs):    A  system
performance check must be made during every  12  hour shift.  For each SPCC
compound in the daily calibration a minimum response factor of 0.050 must
be obtained.  This is the  same  check that is applied  during  the initial
calibration.  If the  minimum response factors are not met, the system must
be evaluated, and corrective action must be taken before  sample analysis
begins.   The minimum RF  for  semivolatile SPCCs is  0.050.   Some possible
problems  are   standard   mixture  degradation,  injection   port  inlet
contamination, contamination at  the  front end  of the  analytical  column,
                            8270B - 14                        Revision 2
                                                          September 1994

-------
and active sites in  the column or chromatographic system.  This check must
be met before analysis begins.

      7.4.4  Calibration  Check  Compounds  (CCCs):    After  the  system
performance  check is  met,  CCCs  listed in Table 4 are  used  to  check the
validity of  the initial calibration.

      Calculate the percent drift using:

                              C,  -  Cc
                    % Drift = -  x 100
where:

      C| =    Calibration  Check Compound  standard concentration.
      Cc =   Measured  concentration using selected quantisation method.

      If the percent difference  for each CCC is less  than or equal to 20%,
the initial calibration  is assumed to be valid.  If the criterion is not
met (>  20%  drift)  for  any one  CCC,  corrective  action must  be taken.
Problems similar to those listed under SPCCs could affect this criterion.
If no source of the problem can  be determined after corrective action has
been  taken, a  new  five-point   calibration  must  be  generated.    This
criterion must  be met before sample analysis begins.  If the CCCs are not
analytes required by the  permit,  then all required analytes must meet the
20% drift criterion.

      7.4.5  The  internal standard responses and retention times  in the
calibration check standard must  be evaluated immediately after or during
data acquisition.   If  the retention time for any internal standard changes
by more than 30 seconds  from the last  calibration check  (12  hours), the
chromatographic system must be inspected for malfunctions and corrections
must  be  made,  as required.   If the  EICP  area  for  any of the  internal
standards changes by a factor of two  (-50% to  +100%) from the last daily
calibration check  standard, the  mass spectrometer must be  inspected for
malfunctions  and  corrections   must  be  made,  as  appropriate.    When
corrections are made,  reanalysis of samples analyzed while the system was
malfunctioning  is required.

7.5   GC/MS analysis

      7.5.1  It  is  highly recommended that the extract be  screened  on  a
GC/FID  or  GC/PID  using  the  same type  of  capillary column.   This  will
minimize  contamination  of  the  GC/MS  system  from  unexpectedly  high
concentrations  of organic compounds.

      7.5.2  Spike  the  1  ml extract obtained from sample preparation with
10 n\- of the internal  standard  solution just prior to  analysis.

      7.5.3 Analyze the  1 ml extract by GC/MS using a 30 m x 0.25 mm (or
0.32 mm) silicone-coated  fused-silica capillary column.  The volume to be
injected should ideally contain  100 ng of base/neutral  and 200 ng of acid


                            8270B -  15                         Revision  2
                                                          September 1994

-------
surrogates  (for  a  1  /il_  injection).    The  recommended GC/MS  operating
conditions to be used are specified in Sec.  7.3.

      7.5.4  If the  response for any quantitation ion exceeds the initial
calibration curve range of the GC/MS  system,  extract  dilution  must take
place.  Additional  internal  standard must be added to the diluted extract
to  maintain  the  required 40  ng//uL   of  each   internal  standard  in  the
extracted volume.  The diluted extract must be reanalyzed.

      7.5.5  Perform all  qualitative  and quantitative measurements  as
described in Sec. 7.6.  Store the extracts at 4°C,  protected from  light
in screw-cap vials equipped with unpierced Teflon lined septa.

7.6   Data interpretation

      7.6.1  Qualitative analysis

             7.6.1.1     The  qualitative  identification   of  compounds
      determined  by  this  method  is  based on retention  time,  and  on
      comparison of the  sample mass spectrum, after background correction,
      with  characteristic  ions   in   a  reference  mass spectrum.    The
      reference mass  spectrum must be generated by the laboratory using
      the conditions  of this  method.   The characteristic  ions  from the
      reference mass spectrum are defined to be the  three ions of greatest
      relative intensity, or any  ions  over 30% relative intensity if less
      than three  such ions  occur in  the reference  spectrum.   Compounds
      should be identified as present when the criteria below are met.

                  7.6.1.1.1    The intensities  of the characteristic ions
             of a compound maximize in  the same scan  or within one scan of
             each  other.   Selection of  a peak  by  a data  system target
             compound  search  routine  where  the search  is based  on  the
             presence  of a target  chromatographic   peak containing  ions
             specific  for  the  target  compound  at  a  compound-specific
             retention time will be accepted as  meeting this  criterion.

                  7.6.1.1.2    The RRT of the  sample component  is within
             ± 0.06  RRT  units of the RRT of the  standard component.

                  7.6.1.1.3    The   relative     intensities     of    the
             characteristic   ions   agree  within 30%   of   the   relative
             intensities   of   these  ions   in   the   reference  spectrum.
             (Example:    For  an  ion   with  an   abundance  of  50%  in  the
             reference spectrum,  the corresponding  abundance in a sample
             spectrum  can  range between  20% and  80%.)

                  7.6.1.1.4    Structural isomers that produce very similar
             mass  spectra  should  be identified  as  individual  isomers if
             they   have   sufficiently   different   GC   retention  times.
             Sufficient  GC resolution  is  achieved  if  the  height  of the
             valley  between two isomer peaks is less than  25% of the sum
             of the  two  peak  heights.   Otherwise,  structural isomers are
             identified  as isomeric pairs.


                            8270B - 16                        Revision 2
                                                          September 1994

-------
            7.6.1.1.5    Identification  is  hampered  when  sample
      components are not  resolved  chromatographically and produce
      mass  spectra  containing ions contributed  by more  than  one
      analyte.  When gas chromatographic peaks obviously represent
      more  than one sample  component  (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of analyte  spectra  and  background
      spectra  is important.   Examination  of  extracted ion current
      profiles  of  appropriate  ions  can aid  in  the  selection  of
      spectra,  and  in  qualitative  identification of  compounds.
      When analytes coelute (i.e.,  only one chromatographic peak is
      apparent), the identification criteria  can  be met,  but each
      analyte  spectrum will contain extraneous ions contributed by
      the coeluting compound.

      7.6.1.2     For  samples containing components not associated
with the calibration standards, a library search may be made for the
purpose of tentative identification.  The necessity to perform this
type of  identification  will be determined  by the  purpose of  the
analyses  being  conducted.    Computer  generated  library  search
routines  should  not   use  normalization  routines   that   would
misrepresent the library  or  unknown spectra when  compared to each
other.   For  example, the RCRA permit or waste delisting requirements
may require the reporting  of nontarget analytes.  Only after visual
comparison of sample spectra with the nearest library searches will
the  mass  spectral   interpretation  specialist  assign  a  tentative
identification. Guidelines for making tentative identification are:

      (1)   Relative  intensities of  major ions  in the  reference
spectrum (ions > 10% of  the  most abundant ion) should be present in
the sample spectrum.

      (2)   The relative intensities of the major ions  should  agree
within  + 20%.   (Example:  For an ion with  an abundance of 50% in the
standard spectrum,  the corresponding  sample  ion  abundance must  be
between 30 and 70%.)

      (3)   Molecular ions present  in the reference spectrum should
be present in the sample spectrum.

      (4)   Ions  present  in  the sample  spectrum  but  not in  the
reference  spectrum  should  be  reviewed for  possible  background
contamination or presence of coeluting compounds.

      (5)   Ions present  in  the reference spectrum  but  not in  the
sample  spectrum should be reviewed for  possible subtraction from the
sample  spectrum because  of background contamination  or  coeluting
peaks.   Data system library  reduction programs can sometimes create
these discrepancies.
                      8270B  -  17                         Revision  2
                                                    September  1994

-------
7.6.2 Quantitative analysis

      7.6.2.1     When  a  compound   has  been   identified,   the
quantitation  of that  compound  will  be based  on the  integrated
abundance from the EICP of the primary characteristic ion.

      7.6.2.2     If the  %RSD of  a  compound's relative  response
factor is 15% or less,  then the concentration in the extract may be
determined  using the  average response  factor  (RF)  from  initial
calibration data (7.4.5.2) and the following equation:.

                      (Ax  x CJ
      Cex (mg/L)  =
                      (Ais  x  RF)

where Cex  is  the  concentration of  the  compound  in  the extract,  and
the other terms are as defined in Sec. 7.4.3.

      7.6.2.3     Alternatively,  the regression line fitted to  the
initial  calibration (Sec.  7.3.5.1) may be used for determination of
the extract concentration.

      7.6.2.4     Compute  the concentration of the  analyte in  the
sample using the equations in Sees.  7.6.2.4.1 and 7.6.2.4.2.

            7.6.2.4.1    The   concentration  of the  analyte  in  the
      liquid   phase   of  the  sample  is  calculated   using   the
      concentration of the analyte in  the extract and the volume of
      liquid extracted, as follows:
            Concentration  in  liquid  (Mg/L)  = (Cm x Vex)
                                                V
      where:

            Vex    =     extract volume,  in ml
            V0     =     volume of liquid extracted,  in L.

            7.6.2.4.2    The concentration  of the  analyte  in  the
      solid   phase  of  the   sample   is  calculated   using   the
      concentration of the pollutant in the extract and the weight
      of the  solids, as follows:

            Concentration  in  solid  (M9/kg) = iCex_x_Vexl
      where:

            Vex    =     extract volume, in ml
            Ws     =     sample weight, in kg.

      7.6.2.5     Where applicable, an estimate of concentration for
noncalibrated components in the sample should be made.  The formulae

                      8270B - 18                        Revision 2
                                                    September 1994

-------
            given  above  should  be  used with the following modifications:  The
            areas  Ax and A.  should be  from the  total  ion chromatograms and the
            RF  for the  compound should be assumed to be 1.  The concentration
            obtained  should be reported  indicating  (1) that the  value  is an
            estimate  and  (2)  which internal standard  was used  to determine
            concentration.     Use   the  nearest  internal   standard  free  of
            interferences.

                   7.6.2.6     Quantitation  of  multicomponent compounds  (e.g.
            Aroclors)   is   beyond   the  scope   of  Method  8270.     Normally,
            quantitation is performed  using  a GC/ECD  by Method 8081.


8.0   QUALITY CONTROL

      8.1   Each  laboratory that  uses  these methods  is required  to operate a
formal quality control program.  The minimum requirements of this program consist
of an initial  demonstration of laboratory capability  and an ongoing analysis of
spiked  samples  to evaluate  and document  quality  data.   The laboratory must
maintain records to document  the quality of the data generated.   Ongoing data
quality checks are compared with established performance criteria to determine
if the results of analyses meet the performance characteristics  of the method.
When results of sample  spikes  indicate atypical  method  performance, a quality
control reference  sample  (Sec. 8.5.1)  must be  analyzed  to confirm  that  the
measurements were performed in an in-control mode of operation.

      8.2   Before  processing  any  samples, the  analyst   should  demonstrate,
through the analysis of a method blank, that interferences from  the analytical
system, glassware, and reagents are under control.  Each time a  set of samples
is extracted or there is a change in  reagents, a method blank should be processed
as a safeguard against chronic laboratory contamination.  The blanks should be
carried through all stages of sample preparation and measurement.

      8.3   The  experience  of  the  analyst  performing  GC/MS  analyses  is
invaluable to  the  success of the methods.   Each day that analysis is performed,
the  daily  calibration   standard  should  be  evaluated  to determine  if  the
chromatographic system is  operating properly.   Questions  that should  be  asked
are:   Do  the  peaks look normal?;  Is  the response obtained comparable to  the
response  from  previous  calibrations?    Careful  examination  of the  standard
chromatogram can  indicate  whether  the column  is  still  good, the  injector is
leaking, the injector septum needs  replacing, etc.  If any changes are made to
the system (e.g. column changed),  recalibration of the system must take place.

      8.4   Required instrument QC  is  found  in the following sections

            8.4.1  The   GC/MS   system  must  be   tuned   to  meet   the   DFTPP
      specifications in Sees. 7.3.1 and 7.4.1.

            8.4.2  There  must  be an  initial  calibration  of the GC/MS system as
      specified in Sec. 7.3.

            8.4.3  The GC/MS system  must meet the SPCC criteria specified in Sec.
      7.4.3 and the CCC criteria in Sec. 7.4.4,  each 12 hours.


                                  8270B - 19                        Revision 2
                                                                September 1994

-------
      8.5   To  establish  the ability  to  generate  acceptable  accuracy  and
precision, the analyst must perform the following operations.

            8.5.1 A  quality  control   (QC)   reference sample  concentrate  is
      required containing each base/neutral  analyte  at  a  concentration of 100
      mg/L and each acid analyte  at a  concentration  of  200  mg/L  in acetone or
      methanol.  (See Sec.  5.5.1  of Method 3500 for minimum requirements.)  The
      QC  reference  sample  concentrate  may be  prepared from pure  standard
      materials  or  purchased  as  certified  solutions.    If  prepared by  the
      laboratory, the QC reference sample concentrate must be made using stock
      standards prepared independently from those used for calibration.

            8.5.2 Using a pi pet,  prepare QC reference samples at a concentration
      of 100 jug/L by adding 1.00 ml of QC reference sample concentrate to each
      of four 1-L aliquots of water.

            8.5.3 Analyze the well-mixed QC reference samples according to the
      method beginning in Sec. 7.1 with extraction of the samples.

            8.5.4 Calculate the average recovery (x) in  jiig/L,  and the standard
      deviation of the recovery  (s) in ng/L,  for each  analyte of interest using
      the four results.

            8.5.5 For  each analyte  compare s  and x with  the  corresponding
      acceptance criteria_for precision  and accuracy, respectively,  found in
      Table 6.   If s  and x  for all analytes  meet  the acceptance  criteria,  the
      system performance is  acceptable and analysis of actual samples can_begin.
      If any individual s exceeds the precision limit  or any individual x falls
      outside  the  range   for   accuracy,   then   the  system   performance  is
      unacceptable for that analyte.

            NOTE: The  large number of analytes  in  Table 6 present a substantial
                  probability that one or more will  fail at  least  one of the
                  acceptance  criteria  when  all analytes  of  a  given method are
                  analyzed.

            8.5.6 When one  or more of the analytes tested fail at least one of
      the  acceptance   criteria,  the  analyst must proceed  according  to  Sec.
      8.5.6.1 or 8.5.6.2.

                  8.5.6.1     Locate  and correct  the source  of the problem and
            repeat the test for  all  analytes of  interest beginning with Sec.
            8.5.2.

                  8.5.6.2     Beginning with Sec. 8.5.2,  repeat  the test only
            for those analytes that failed to meet criteria.  Repeated failure,
            however, will confirm a general problem with the measurement system.
            If this  occurs,  locate and correct the  source  of the problem and
            repeat the test for  all compounds  of interest beginning with Sec.
            8.5.2.

      8.6   The  laboratory  must,  on an  ongoing basis, analyze a  method blank,
a matrix spike,  and a  replicate for each analytical batch  (up to a maximum of 20


                                  8270B - 20                        Revision 2
                                                                September 1994

-------
samples/batch) to assess accuracy.   For  soil and waste samples where detectable
amounts of organics are present, replicate samples may be appropriate in place
of matrix  spiked  samples.   For laboratories analyzing one  to  ten samples per
month, at least one spiked sample per month is required.

            8.6.1  The  concentration  of the  spike  in  the  sample should  be
      determined as follows:

                   8.6.1.1     If, as in compliance monitoring, the concentration
            of  a  specific analyte  in the sample  is being checked  against  a
            regulatory concentration  limit, the  spike  should  be at that limit
            or 1 to 5 times higher than  the background concentration determined
            in Sec. 8.6.2, whichever  concentration would be larger.

                   8.6.1.2     If the  concentration of a  specific  analyte  in  a
            water  sample is not being checked against a limit specific to that
            analyte, the spike should be at 100 jug/L or 1  to  5 times higher than
            the background concentration  determined in  Step  8.6.2,  whichever
            concentration  would be  larger.   For other  matrices,  recommended
            spiking concentration is  20 times the EQL.

                   8.6.1.3     If  it  is  impractical   to  determine  background
            levels before spiking (e.g. maximum holding times will be exceeded),
            the   spike   concentration  should   be   at  (1)   the  regulatory
            concentration  limit, if any; or,  if  none (2)  the  larger of either
            5 times  higher than the  expected background concentration  or 100
            jug/L.  For other matrices, recommended spiking concentration is 20
            times  the EQL.

            8.6.2  Analyze  one  sample  aliquot   to   determine  the  background
      concentration  (B)  of  each analyte.   If  necessary,  prepare  a  new  QC
      reference sample concentrate  (Sec. 8.5.1)  appropriate for the background
      concentration in the sample.   Spike  a second sample aliquot with 1.00 ml
      of the  QC reference  sample concentrate  and analyze it  to determine the
      concentration after spiking (A) of each  analyte.   Calculate each percent
      recovery (p) as  100(A-B)%/T, where T is the  known true  value of the spike.

            8.6.3  Compare  the percent recovery (p) for each analyte in a water
      sample with  the corresponding QC  acceptance criteria found  in  Table  6.
      These acceptance  criteria were calculated  to  include an  allowance for
      error in  measurement of  both  the background  and  spike  concentrations,
      assuming a spike to background ratio of 5:1.  This error will be accounted
      for to the extent  that  the analyst's spike to background ratio approaches
      5:1.   If spiking was performed at a concentration lower than 100 /^g/L, the
      analyst must use either the QC acceptance  criteria  presented in Table 6,
      or optional  QC acceptance  criteria  calculated for  the  specific  spike
      concentration.  To calculate optional  acceptance criteria for the recovery
      of an analyte:  (1)  Calculate  accuracy (x') using the equation  found  in
      Table 7, substituting  the spike concentration  (T)  for C;  (2)  calculate
      overall  precision  (S')  using the equation in Table 7,  substituting x' for
      x;  (3)  calculate  the range for recovery  at the spike  concentration  as
      (lOOx'/T)  + 2.44(100S'/T)%.
                                  8270B -  21                         Revision 2
                                                                September 1994

-------
            8.6.4  If  any individual  p falls outside  the  designated range for
      recovery,  that  analyte  has  failed  the  acceptance  criteria.    A  check
      standard containing each analyte that failed the criteria must be analyzed
      as described in Sec. 8.7.

      8.7   If  any analyte  in  a  sample  fails  the  acceptance  criteria  for
recovery in Sec.  8.6,  a  QC reference  sample containing each analyte that failed
must be prepared and analyzed.

      NOTE: The  frequency for the required analysis  of a QC reference sample
            will depend upon the  number of analytes  being simultaneously tested,
            the  complexity of the sample  matrix,  and the performance of the
            laboratory.   If the  entire  list of analytes in Table  6  must be
            measured  in  the  sample  in  Sec.  8.6,  the  probability that  the
            analysis of a QC reference sample will be required is high.  In this
            case, the QC reference sample should  be routinely  analyzed with the
            spiked sample.

            8.7.1  Prepare the  QC reference sample by adding  1.0  ml  of the QC
      reference sample concentrate (Sec. 8.5.1 or 8.6.2)  to 1 L of water.  The
      QC  reference sample  needs  only  to  contain the  analytes that  failed
      criteria in the test in Sec. 8.6.

            8.7.2  Analyze the QC reference sample to determine the concentration
      measured  (A)  of each analyte.   Calculate  each  percent recovery (ps) as
      100(A/T)%, where T  is the  true value of the standard concentration.

            8.7.3  Compare the  percent  recovery (pj  for each analyte with the
      corresponding QC acceptance criteria found  in Table 6.  Only  analytes that
      failed the test in  Sec. 8.6 need to be compared with these criteria.  If
      the recovery of any such analyte falls outside the designated range, the
      laboratory performance for that analyte  is judged to be out of control,
      and  the problem  must  be  immediately  identified   and  corrected.    The
      analytical result for that analyte in the unspiked  sample is suspect and
      may not be reported for regulatory compliance purposes.

      8.8   As part of  the QC  program  for the  laboratory, method accuracy for
each matrix studied must be assessed  and  records  must  be maintained.  After the
analysis of five spiked samples _(of the same matrix)  as in Sec.  8.6, calculate
the  average  percent  recovery  (p)  and  the  standard  deviation  of  the  percent
recovery (sp).   Express  the accuracy  assessment as  a percent recovery interval
from p  - 2sp to p  +  2sp.   If p = 90% and sp = 10%,  for example,  the accuracy
interval is  expressed as  70-110%.   Update  the  accuracy  assessment  for each
analyte  on  a  regular   basis  (e.g.  after  each  five  to  ten  new  accuracy
measurements).

      8.9   The following procedure should be performed to determine acceptable
accuracy and precision limits for surrogate standards.

            8.9.1  For each sample analyzed,  calculate the percent recovery of
      each surrogate  in the sample.
                                  8270B - 22                        Revision 2
                                                                September 1994

-------
             8.9.2  Once a minimum of thirty samples of the  same matrix have been
      analyzed,  calculate  the  average  percent   recovery   (P)  and  standard
      deviation of the  percent recovery  (s) for each of the  surrogates.

             8.9.3  For a given matrix,  calculate  the upper  and  lower control
      limit  for method performance for each surrogate standard.  This should be
      done as follows:

             Upper  Control  Limit  (UCL)  =  P  + 3s
             Lower  Control  Limit  (LCL)  =  P  - 3s

             8.9.4  For aqueous and soil matrices,  these laboratory-established
      surrogate  control limits  should,  if applicable,  be   compared  with  the
      control limits listed in Table 8.  The limits given  in  Table 8 are multi-
      laboratory  performance-based  limits for  soil  and  aqueous  samples,  and
      therefore, the  single-laboratory limits established in Sec.  8.9.3 must
      fall within those given in Table 8 for these matrices.

             8.9.5  If  recovery is not within limits, the following procedures are
      required.

             •      Check to  be  sure   there  are  no   errors  in  calculations,
                   surrogate  solutions  and internal   standards.   Also,  check
                   instrument  performance.

             •      Recalculate the data and/or reanalyze the  extract if any of
                   the above checks  reveal  a problem.

             •      Reextract and  reanalyze  the sample if none of the above are
                   a problem or  flag the  data  as "estimated concentration".

             8.9.6  At  a minimum,  each laboratory should update  surrogate recovery
      limits on a matrix-by-matrix basis,  annually.

      8.10   It  is recommended  that the  laboratory  adopt additional  quality
assurance practices for use with this method.   The specific  practices that are
most productive depend  upon the needs  of the  laboratory and  the  nature of the
samples.   Field duplicates may be analyzed  to  assess  the  precision  of  the
environmental measurements. When doubt exists over the identification of a peak
on the chromatogram,  confirmatory techniques such  as  gas chromatography with a
dissimilar column, specific element  detector, or  a  mass  spectrometer must be
used.   Whenever possible,  the laboratory  should analyze standard  reference
materials and participate in relevant performance  evaluation studies.


9.0   METHOD PERFORMANCE

      9.1    Method 8250 (the  packed column version of Method 8270)  was tested
by 15 laboratories using  organic-free reagent water,  drinking water,  surface
water, and industrial  wastewaters spiked at six concentrations over the range 5-
1,300 jLtg/L.  Single operator  accuracy  and  precision,  and  method  accuracy were
found to be directly related to the concentration of the  analyte and essentially
                                  8270B - 23                        Revision 2
                                                                September 1994

-------
independent  of  the   sample   matrix.   Linear   equations   to  describe  these
relationships are presented in Table 7.

      9.2    Chromatograms from calibration standards analyzed with Day 0 and Day
7 samples  were  compared to  detect  possible deterioration  of  GC performance.
These recoveries (using Method 3510 extraction) are presented in Table 9.

      9.3    Method  performance  data  (using  Method  3541  Automated  Soxhlet
extraction) are presented in Table 10.  Single laboratory accuracy and precision
data were  obtained  for semivolatile organics  in  a clay soil by spiking at a
concentration of 6 mg/kg  for each compound.  The spiking solution  was mixed into
the soil during addition and  then  allowed  to  equilibrate for approximately 1 hr
prior to extraction.   The spiked samples were then extracted  by  Method 3541
(Automated Soxhlet).  Three determinations were performed and each extract was
analyzed by gas chromatography/ mass  spectrometry following Method 8270. The low
recovery of the more volatile compounds is probably due to volatilization  losses
during equilibration.  These  data are listed  in Table 11  and  were taken from
Reference 9.
10.0  REFERENCES

1.    U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test  Procedures for the
      Analysis of Pollutants Under the Clean Water Act,  Method 625," October 26,
      1984.

2.    U.S.  EPA Contract  Laboratory Program,  Statement  of  Work  for  Organic
      Analysis, July  1985,  Revision.

3.    Eichelberger, J.W., L.E. Harris,  and  W.L.  Budde, "Reference Compound to
      Calibrate   Ion   Abundance   Measurement   in  Gas   Chromatography-Mass
      Spectrometry Systems," Analytical Chemistry, 47, 995-1000,  1975.

4.    "Method Detection Limit for Methods  624 and  625," Olynyk, P., W.L. Budde,
      and J.W. Eichelberger, Unpublished report, October  1980.

5.    "Inter!aboratory Method Study for EPA Method 625-Base/Neutrals, Acids, and
      Pesticides," Final Report for EPA Contract 68-03-3102  (in preparation).

6.    Burke,  J.A. "Gas  Chromatography  for  Pesticide  Residue Analysis;  Some
      Practical  Aspects,"  Journal  of  the  Association of Official  Analytical
      Chemists, 48,  1037, 1965.

7.    Lucas, S.V.; Kornfeld,  R.A.  "GC-MS  Suitability Testing of RCRA Appendix
      VIII and Michigan List Analytes "; U.S. Environmental  Protection Agency,
      Environmental  Monitoring  and Support  Laboratory,  Cincinnati,  OH 45268,
      February 20, 1987, Contract  No. 68-03-3224.

8.    Engel, T.M.; Kornfeld,  R.A.; Warner,  J.S.;  Andrews, K.D.   "Screening of
      Semivolatile Organic  Compounds  for  Extractability and Aqueous Stability
      by  SW-846,     Method  3510";  U.S.   Environmental  Protection  Agency,
      Environmental  Monitoring  and Support  Laboratory,  Cincinnati,  OH 45268,
      June 5,  1987, Contract 68-03-3224.


                                   8270B - 24                        Revision  2
                                                                September 1994

-------
9.    Lopez-Avila, V.  (W.  Beckert,  Project  Officer);  "Development of a Soxtec
      Extraction Procedure for  Extraction of  Organic  Compounds from Soils and
      Sediments";    U.S.  Environmental  Protection  Agency.    Environmental
      Monitoring  and  Support Laboratory.   Las Vegas,  NV,  October  1991;  EPA
      600/X-91/140.
                                  8270B - 25                        Revision 2
                                                                September 1994

-------
                                   TABLE 1.
                CHARACTERISTIC IONS FOR SEMIVOLATILE COMPOUNDS
Compound
Retention    Primary   Secondary
Time (min.)  Ion       Ion(s)
2-Picoline
Aniline
Phenol
Bis(2-chloroethyl) ether
2-Chlorophenol
1,3-Dichlorobenzene
l,4-Dichlorobenzene-d4 (I.S.)
1,4-Dichlorobenzene
Benzyl alcohol
1,2-Dichlorobenzene
N-Ni trosomethylethyl ami ne
Bis(2-chloroisopropyl) ether
Ethyl  carbamate
Thiophenol (Benzenethiol)
Methyl methanesulfonate
N-Nitrosodi-n-propylamine
Hexachloroethane
Maleic anhydride
Nitrobenzene
Isophorone
N-Nitrosodiethyl amine
2-Nitrophenol
2,4-Dimethylphenol
p-Benzoquinone
Bis(2-chloroethoxy)methane
Benzoic acid
2,4-Dichlorophenol
Trimethyl phosphate
Ethyl  methanesulfonate
1,2,4-Trichlorobenzene
Naphthalene-d8 (I.S.)
Naphthalene
Hexachlorobutadiene
Tetraethyl pyrophosphate
Diethyl sulfate
4-Chloro-3-methylphenol
2-Methylnaphthalene
2-Methylphenol
Hexachloropropene
Hexachlorocyclopentadi ene
N-Nitrosopyrrolidine
Acetophenone
4-Methylphenol
2,4,6-Trichlorophenol
o-Toluidine
3-Methylphenol
     3.75'
     5.68
     5.77
     5.82
     5.97
     6.27
     6.35
     6.40
     6.78
     6.
     6.
     7.
     7.
     7.
     7.
     7.
     7,
     7.
     7,
  85
  97
  22
  27
  42
  48
  55
  65
  65
  87
     8.53
     8.70
     8.75
      .03
      ,13
9.
9.
9.23
9.
9.
9.
9.
      .38
      .48
      ,53
      .62
     9.67
     9.75
     9.82
    10.43
    11.07
    11.37
    11.68
    11.87
    12.40
    12.45
    12.60
    12.65
    12.67
    12.82
    12.85
    12.87
    12.93
 93     66,92
 93     66,65
 94     65,66
 93     63,95
128     64,130
146     148,111
152     150,115
146     148,111
108     79,77
146     148,111
 88     42,88,43,56
 45     77,121
 62     62,44,45,74
110     110,66,109,84
 80     80,79,65,95
 70     42,101,130
117     201,199
 54     54,98,53,44
 77     123,65
 82     95,138
102     102,42,57,44,56
139     109,65
122     107,121
108     54,108,82,80
 93     95,123
122     105,77
162     164,98
110     110,79,95,109,140
 79     79,109,97,45,65
180     182,145
136     68
128     129,127
225     223,227
 99     99,155,127,81,109
139     139,45,59,99,111,125
107     144,142
142     141
107     107,108,77,79,90
213     213,211,215,117,106,141
237     235,272
100     100,41,42,68,69
105     71,105,51,120
107     107,108,77,79,90
196     198,200
106     106,107,77,51,79
107     107,108,77,79,90
                                  8270B  - 26
                                  Revision 2
                              September 1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
  Retention
  Time  (min.)
Primary
Ion
Secondary
Ion(s)
2-Chloronaphthalene
N-Nitrosopiperi dine
1,4-Phenylenedi ami ne
1-Chloronaphthalene
2-Nitroaniline
5-Chloro-2-methylaniline
Dimethyl phthalate
Acenaphthylene
2,6-Dinitrotoluene
Phthalic anhydride
o-Anisidine
3-Nitroaniline
Acenaphthene-d10  (I.S.)
Acenaphthene
2,4-Dinitrophenol
2,6-Dinitrophenol
4-Chloroaniline
Isosafrole
Dibenzofuran
2,4-Diaminotoluene
2,4-Dinitrotoluene
4-Nitrophenol
2-Naphthylamine
1,4-Naphthoquinone
p-Cresidine
Dichlorovos
Diethyl phthalate
Fluorene
2,4,5-Trimethylaniline
N-Ni trosodi butyl ami ne
4-Chlorophenyl phenyl ether
Hydroquinone
4,6-Dinitro-2-methylphenol
Resorcinol
N-Nitrosodiphenylamine
Safrole
Hexamethyl phosphoramide
3-(Chloromethyl)pyridine hydrochl
Diphenylamine
1,2,4,5-Tetrachlorobenzene
1-Naphthylamine
l-Acetyl-2-thiourea
4-Bromophenyl phenyl ether
Toluene diisocyanate
2,4,5-Trichlorophenol
Hexachlorobenzene
     13.30      162      127,164
     13.55      114      42,114,55,56,41
     13.62      108      108,80,53,54,52
     13.65a     162      127,164
     13.75       65      92,138
     14.28      106      106,141,140,77,89
     14.48      163      194,164
     14.57      152      151,153
     14.62      165      63,89
     14.62      104      104,76,50,148
     15.00      108      80,108,123,52
     15.02      138      108,92
     15.05      164      162,160
     15.13      154      153,152
     15.35      184      63,154
     15.47      162      162,164,126,98,63
     15.50      127      127,129,65,92
     15.60      162      162,131,104,77,51
     15.63      168      139
     15.78      121      121,122,94,77,104
     15.80      165      63,89
     15.80      139      109,65
     16.00a     143      115,116
     16.23      158      158,104,102,76,50,130
     16.45      122      122,94,137,77,93
     16.48      109      109,185,79,145
     16.70      149      177,150
     16.70      166      165,167
     16.70      120      120,135,134,91,77
     16.73       84      84,57,41,116,158
     16.78      204      206,141
     16.93      110      110,81,53,55
     17.05      198      51,105
     17.13      110      110,81,82,53,69
     17.17      169      168,167
     17.23      162      162,162,104,77,103,135
     17.33      135      135,44,179,92,42
oride!7.50       92      92,127,129,65,39
     17.54a     169      168,167
     17.97      216      216,214,179,108,143,218
     18.20      143      143,115,89,63
     18.22      118      43,118,42,76
     18.27      248      250,141
     18.42      174      174,145,173,146,132,91
     18.47      196      196,198,97,132,99
     18.65      284      142,249
                                  8270B - 27
                                   Revision 2
                               September 1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
Nicotine
Pentachlorophenol
5-Nitro-o-toluidine
Thionazine
4-Nitroaniline
Phenanthrene-d10(i .s.)
Phenanthrene
Anthracene
1,4-Dinitrobenzene
Mevinphos
Naled
1,3-Dinitrobenzene
Diallate (cis or trans)
1,2-Dinitrobenzene
Diallate (trans or cis)
Pentachlorobenzene
5-Nitro-o-anisidine
Pentachloronitrobenzene
4-Nitroquinoline-l-oxide
Di-n-butyl phthalate
2,3,4,6-Tetrachlorophenol
Dihydrosaffrole
Demeton-0
Fluoranthene
1,3,5-Trinitrobenzene
Dicrotophos
Benzidine
Trifluralin
Bromoxynil
Pyrene
Monocrotophos
Phorate
Sulfall ate
Demeton-S
Phenacetin
Dimethoate
Phenobarbital
Carbofuran
Octamethyl pyrophosphoramide
4-Aminobiphenyl
Dioxathion
Terbufos
a,a-Dimethylphenylamine
Pronamide
Aminoazobenzene
Dichlone
    18.70      84      84,133,161,162
    19.25     266      264,268
    19.27     152     77,152,79,106,94
    19.35     107     96,107,97,143,79,68
    19.37     138     138,65,108,92,80,39
    19.55     188     94,80
    19.62     178     179,176
    19.77     178     176,179
    19.83     168     168,75,50,76,92,122
    19.90     127     127,192,109,67,164
    20.03     109     109,145,147,301,79,189
    20.18     168     168,76,50,75,92,122
    20.57      86     86,234,43,70
    20.58     168     168,50,63,74
    20.78      86     86,234,43,70
    21.35     250     250,252,108,248,215,254
    21.50     168     168,79,52,138,153,77
    21.72     237     237,142,214,249,295,265
    21.73     174     174,101,128,75,116
    21.78     149     150,104
    21.88     232     232,131,230,166,234,168
    22.42     135     135,64,77
    22.72      88     88,89,60,61,115,171
    23.33     202     101,203
    23.68      75     75,74,213,120,91,63
    23.82     127     127,67,72,109,193,237
    23.87     184     92,185
    23.88     306     306,43,264,41,290
    23.90     277     277,279,88,275,168
    24.02     202     200,203
    24.08     127     127,192,67,97,109
    24.10      75     75,121,97,93,260
    24.23     188     188,88,72,60,44
    24.30      88     88,60,81,89,114,115
    24.33     108     180,179,109,137,80
    24.70      87     87,93,125,143,229
    24.70     204     204,117,232,146,161
    24.90     164     164,149,131,122
    24.95     135     135,44,199,286,153,243
    25.08     169     169,168,170,115
    25.25      97     97,125,270,153
    25.35     231     231,57,97,153,103
    25.43      58     58,91,65,134,42
    25.48     173     173,175,145,109,147
    25.72     197     92,197,120,65,77
    25.77     191     191,163,226,228,135,193
                                  8270B - 28
                                  Revision 2
                              September 1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
Dinoseb
Disulfoton
Fluchloralin
Mexacarbate
4,4'-Oxydianiline
Butyl benzyl phthalate
4-Nitrobiphenyl
Phosphamidon
2-Cyclohexyl-4,6-Dinitrophenol
Methyl parathion
Carbaryl
Dimethyl aminoazobenzene
Propylthiouracil
Benz(a)anthracene
Chrysene-d12 (I.S.)
3,3'-Dichlorobenzidine
Chrysene
Malathion
Kepone
Fenthion
Parathion
Anilazine
Bis(2-ethylhexyl) phthalate
3,3'-Dimethylbenzidine
Carbophenothion
5-Nitroacenaphthene
Methapyrilene
Isodrin
Captan
Chlorfenvinphos
Crotoxyphos
Phosmet
EPN
Tetrachlorvinphos
Di-n-octyl phthalate
2-Aminoanthraquinone
Barban
Aramite
Benzo(b)fluoranthene
Nitrofen
Benzo(k)fluoranthene
Chlorobenzilate
Fensulfothion
Ethion
Diethylstilbestrol
Famphur
    25.83     211     211,163,147,117,240
    25.83      88     88,97,89,142,186
    25.88     306     306,63,326,328,264,65
    26.02     165     165,150,134,164,222
    26.08     200     200,108,171,80,65
    26.43     149     91,206
    26.55     199     199,152,141,169,151
    26.85     127     127,264,72,109,138
    26.87     231     231,185,41,193,266
    27.03     109     109,125,263,79,93
    27.17     144     144,115,116,201
    27.50     225     225,120,77,105,148,42
    27.68     170     170,142,114,83
    27.83     228     229,226
    27.88     240     120,236
    27.88     252     254,126
    27.97     228     226,229
    28.08     173     173,125,127,93,158
    28.18     272     272,274,237,178,143,270
    28.37     278     278,125,109,169,153
    28.40     109     109,97,291,139,155
    28.47     239     239,241,143,178,89
    28.47     149     167,279
    28.55     212     212,106,196,180
    28.58     157     157,97,121,342,159,199
    28.73     199     199,152,169,141,115
    28.77      97     97,50,191,71
    28.95     193     193,66,195,263,265,147
    29.47      79     79,149,77,119,117
    29.53     267     267,269,323,325,295
    29.73     127     127,105,193,166
    30.03     160     160,77,93,317,76
    30.11     157     157,169,185,141,323
    30.27     329     109,329,331,79,333
    30.48     149     167,43
    30.63     223     223,167,195
    30.83'    222     222,51,87,224,257,153
    30.92     185     185,191,319,334,197,321
    31.45     252     253,125
    31.48     283     283,285,202,139,253
    31.55     252     253,125
    31.77     251     251,139,253,111,141
    31.87     293     293,97,308,125,292
    32.08     231     231,97,153,125,121
    32.15     268     268,145,107,239,121,159
    32.67     218     218,125,93,109,217
                                  8270B - 29
                                  Revision 2
                              September  1994

-------
                                   TABLE 1.
                                  (Continued)
Compound
Retention
Time (min.)
Primary   Secondary
Ion       Ion(s)
Tri-p-tolyl phosphate"                32.75      368
Benzo(a)pyrene                        32.80      252
Perylene-d12 (I.S.)                    33.05      264
7,12-Dimethylbenz(a)anthracene        33.25      256
5,5-Diphenylhydantoin                 33.40      180
Captafol                              33.47       79
Dinocap                               33.47       69
Methoxychlor                          33.55      227
2-Acetylaminofluorene                 33.58      181
4,4'-Methylenebis(2-chloroaniline)    34.38      231
3,3'-Dimethoxybenzidine               34.47      244
3-Methylcholanthrene                  35.07      268
Phosalone                             35.23      182
Azinphos-methyl                       35.25      160
Leptophos                             35.28      171
Mirex                                 35.43      272
Tris(2,3-dibromopropyl) phosphate     35.68      201
Dibenz(a,j)acridine                   36.40      279
Mestranol                             36.48      277
Coumaphos                             37.08      362
Indeno(l,2,3-cd)pyrene                39.52      276
Dibenz(a,h)anthracene                 39.82      278
Benzo(g,h,i)perylene                  41.43      276
l,2:4,5-Dibenzopyrene                 41.60      302
Strychnine                            45.15      334
Piperonyl sulfoxide                   46.43      162
Hexachlorophene                       47.98      196
Aldrin                                  --         66
Aroclor-1016                            --        222
Aroclor-1221                            --        190
Aroclor-1232                            --        190
Aroclor-1242                            --        222
Aroclor-1248                            --        292
Aroclor-1254                            --        292
Aroclor-1260                            --        360
a-BHC                                   --        183
j3-BHC                                   --        181
£-BHC                                   --        183
7-BHC  (Lindane)                         --        183
4,4'-DDD                                --        235
4,4'-DDE                                --        246
4,4'-DDT                                --        235
Dieldrin                                --         79
1,2-Diphenylhydrazine                   --         77
Endosulfan  I                            --        195
Endosulfan  II                           --        337
                      368,367,107,165,198
                      253,125
                      260,265
                      256,241,239,120
                      180,104,252,223,209
                      79,77,80,107
                      69,41,39
                      227,228,152,114,274,212
                      181,180,223,152
                      231,266,268,140,195
                      244,201,229
                      268,252,253,126,134,113
                      182,184,367,121,379
                      160,132,93,104,105
                      171,377,375,77,155,379
                      272,237,274,270,239,235
                      137,201,119,217,219,199
                      279,280,277,250
                      277,310,174,147,242
                      362,226,210,364,97,109
                      138,227
                      139,279
                      138,277
                      302,151,150,300
                      334,335,333
                      162,135,105,77
                      196,198,209,211,406,408
                      263,220
                      260,292
                      224,260
                      224,260
                      256,292
                      362,326
                      362,326
                      362,394
                      181,109
                      183,109
                      181,109
                      181,109
                      237,165
                      248,176
                      237,165
                      263,279
                      105,182
                      339,341
                      339,341
                                  8270B - 30
                                  Revision 2
                              September 1994

-------
                                   TABLE 1.
                                  (Continued)
                                   Retention     Primary    Secondary
Compound                           Time  (min.)   Ion        Ion(s)
Endosulfan sulfate                      --        272     387,422
Endrin                                  --        263     82,81
Endrin aldehyde                      ,   --         67     345,250
Endrin ketone                           --        317     67,319
2-Fluorobiphenyl (surr.)                --        172     171
2-Fluorophenol  (surr.)                  --        112     64
Heptachlor                              --        100     272,274
Heptachlor epoxide                      --        353     355,351
Nitrobenzene-d5 (surr.)                 --         82     128,54
N-Nitrosodimethylamine                  --         42     74,44
Phenol-d6 (surr.)                        --         99     42,71
Terphenyl-du (surr.)                    --        244     122,212
2,4,6-Tribromophenol (surr.)            --        330     332,141
Toxaphene                               --        159     231,233
I.S.  = internal standard.
surr. = surrogate.
""Estimated retention times.
bSubstitute for the non-specific mixture,  tricresyl  phosphate.
                                  8270B - 31                        Revision 2
                                                                September 1994

-------
                           TABLE 2.
ESTIMATED QUANTITATION LIMITS (EQLs) FOR SEMIVOLATILE ORGANICS

                                       Estimated
                                      Quantitation
                                          Limits"
Ground water
Semivolatiles M9/L
Acenaphthene
Acenaphthylene
Acetophenone
2-Acetyl ami nof 1 uorene
l-Acetyl-2-thiourea
2-Aminoanthraquinone
Aminoazobenzene
4-Aminobiphenyl
Anilazine
o-Anisidine
Anthracene
Aramite
Azinphos-methyl
Barban
Benz(a)anthracene
Benzo(b)fl uoranthene
Benzo ( k) f 1 uoranthene
Benzole acid
Benzo (g,h,i)perylene
Benzo(a)pyrene
p-Benzoquinone
Benzyl alcohol
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
4-bromophenyl phenyl ether
Bromoxynil
Butyl benzyl phthalate
Captafol
Captan
Carbaryl
Carbofuran
Carbophenothion
Chlorfenvinphos
4-Chloroaniline
Chi orobenzi late
5-Chloro-2 -methyl aniline
4-Chloro-3 -methyl phenol
3-(Chloromethyl)pyridine hydrochloride
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Coumaphos
10
10
10
20
1000
20
10
20
100
10
10
20
100
200
10
10
10
50
10
10
10
20
10
10
10
10
10
10
20
50
10
10
10
20
20
10
10
20
100
10
10
10
10
40
Low Soil/Sediment"
M9Ag
660
660
ND
NO
ND
ND
ND
ND
ND
ND
660
ND
ND
ND
660
660
660
3300
660
660
ND
1300
660
660
660
660
ND
660
ND
ND
ND
ND
ND
ND
1300
ND
ND
1300
ND
660
660
660
660
ND
                          8270B - 32                        Revision 2
                                                        September 1994

-------






Semivolatiles
p-Cresidine
Crotoxyphos
2-Cyclohexyl-4,6-dinitrophenol
Demeton-0
Demeton-S
Diallate (cis or trans)
Diallate (trans or cis)
2,4-Diaminotoluene
Dibenz(a,j)acridine
Dibenz( a, h) anthracene
Dibenzofuran
Dibenzo(a,e)pyrene
Di-n-butyl phthalate
Dichlone
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 ,4-Dichlorobenzene
3,3'-Dichlorobenzidine
2,4-Dichlorophenol
2,6-Dichlorophenol
Dichlorovos
Dicrotophos
Diethyl phthalate
Diethylstilbestrol
Diethyl sulfate
Dimethoate
3,3'-Dimethoxybenzidine
Dimethyl ami noazobenzene
7, 12 -Dimethyl benz( a) anthracene
3, 3' -Dimethyl benzi dine
a, a-Dimethyl phenethyl ami ne
2,4-Dimethylphenol
Dimethyl phthalate
1,2-Dinitrobenzene
1,3-Dinitrobenzene
1,4-Dinitrobenzene
4, 6-Dinitro-2-methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Dinocap
Dinoseb
5,5-Diphenylhydantoin
Di-n-octyl phthalate
TABLE 2.
(Continued)



Ground
M9/1
10
20
100
10
10
10
10
20
10
10
10
10
10
NA
10
10
10
20
10
10
10
10
10
20
100
20
100
10
10
10
ND
10
10
40
20
40
50
50
10
10
100
20
20
10


Estimated
Quantitation
Limits8
water Low Soil/Sediment6
M9/kg
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
660
ND
ND
ND
660
660
660
1300
660
ND
ND
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
660
660
ND
ND
ND
3300
3300
660
660
ND
ND
ND
660
8270B - 33
    Revision 2
September 1994

-------






Semi vol at lies
Disulfoton
EPN
Ethion
Ethyl carbamate
Bis(2-ethylhexy1) phthalate
Ethyl methanesulfonate
Famphur
Fensulfothion
Fenthion
Fluchloralin
Fluoranthene
Fluorene
Hexachlorobenzene
Hexachl orobutadi ene
Hexachl orocycl opentadi ene
Hexachl oroethane
Hexachl orophene
Hexachl oropropene
Hexamethyl phosphorami de
Hydroquinone
Indeno(l,2,3-cd)pyrene
Isodrin
Isophorone
Isosafrole
Kepone
Leptophos
Malathion
Maleic anhydride
Mestranol
Methapyrilene
Methoxychlor
3-Methyl chol anthrene
4,4'-Methylenebis(2-ch1oroanil
Methyl methanesulfonate
2-Methyl naphthalene
Methyl parathion
2-Methylphenol
3-Methyl phenol
4-Methylphenol
Mevinphos
Mexacarbate
Mi rex
Monocrotophos
Naled
TABLE 2.
(Continued)
Estimated
Quantitation
Limits"
Ground water Low Soi
M9/L
10
10
10
50
10
20
20
40
10
20
10
10
10
10
10
10
50
10
20
ND
10
20
10
10
20
10
50
NA
20
100
10
10
ine) NA
10
10
10
10
10
10
10
20
10
40
20





l/Sedimentb
MQ/kg
ND
ND
ND
ND
660
ND
ND
ND
ND
ND
660
660
660
660
660
660
ND
ND
ND
ND
660
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
ND
660
ND
660
ND
ND
ND
ND
ND
8270B - 34
    Revision 2
September 1994

-------






Semivolatiles
Naphthalene
1,4-Naphthoquinone
1-Naphthylamine
2-Naphthylamine
Nicotine
5-Nitroacenaphthene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
5-Nitro-o-anisidine
Nitrobenzene
4-Nitrobiphenyl
Nitrofen
2-Nitrophenol
4-Nitrophenol
5-Nitro-o-toluidine
4-Nitroqui no! ine-1 -oxide
N-Ni trosodi butyl ami ne
N-Ni trosodi ethyl ami ne
N-Ni trosodi phenyl ami ne
N-Ni troso-di-n-propyl ami ne
N-Ni trosopi peri dine
N-Nitrosopyrrolidine
Octamethyl pyrophosphoramide
4,4'-Oxydianiline
Parathion
Pentachlorobenzene
Pentachloronitrobenzene
Pentachlorophenol
Phenacetin
Phenanthrene
Phenobarbital
Phenol
1 , 4- Phenyl ened i ami ne
Phorate
Phosalone
Phosmet
Phosphamidon
Phthalic anhydride
2-Picoline
Piperonyl sulfoxide
Pronamide
Propylthiouracil
Pyrene
TABLE 2.
(Continued)
Estimated
Quantitation
Limits8
Ground water Low Soi
M9/L
10
10
10
10
20
10
50
50
20
10
10
10
20
10
50
10
40
10
20
10
10
20
40
200
20
10
10
20
50
20
10
10
10
10
10
100
40
100
100
ND
100
10
100
10





I/Sediment"
M9/kg
660
ND
ND
ND
ND
ND
3300
3300
ND
ND
660
ND
ND
660
3300
ND
ND
ND
ND
660
660
ND
ND
ND
ND
ND
ND
ND
3300
ND
660
ND
660
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
8270B - 35
    Revision 2
September 1994

-------
                                   TABLE 2.
                                  (Continued)
                                               Estimated
                                              Quantitation
                                                 Limits"
                                      Ground water   Low Soil/Sediment
Semivolatiles                            M9A
Pyridine
Resorcinol
Safrole
Strychnine
Sul fall ate
Terbufos
1 , 2 , 4, 5-Tetrachl orobenzene
2,3,4,6-Tetrachlorophenol
Tetrachlorvinphos
Tetraethyl pyrophosphate
Thionazine
Thiophenol (Benzenethiol )
Toluene diisocyanate
o-Toluidine
1, 2, 4-Trichl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
Trifluralin
2,4,5-Trimethylanil ine
Trimethyl phosphate
1, 3, 5-Tri nitrobenzene
Tris(2,3-dibromopropyl ) phosphate
Tri-p-tolyl phosphate(h)
0,0,0-Triethyl phosphorothioate
ND
100
10
40
10
20
10
10
20
40
20
20
100
10
10
10
10
10
10
10
10
200
10
NT
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
660
660
660
ND
ND
ND
ND
ND
ND
ND
a Sample EQLs are highly matrix-dependent.   The  EQLs listed herein are provided
  for guidance and may not always be achievable.
b EQLs  listed  for soil/sediment are based  on  wet weight.   Normally  data are
  reported on  a  dry weight basis, therefore,  EQLs will  be higher based on the
  % dry weight of each  sample.   These  EQLs  are  based  on a 30 g sample and gel
  permeation chromatography cleanup.
ND = Not determined..
NA = Not applicable.
NT = Not tested.
Other Matrices                                       Factor0

High-concentration soil and sludges by sonicator7~1>
Non-water miscible waste                             75

°EQL  =  (EQL  for Low Soil/Sediment given above in Table 2) X (Factor).


                                  8270B - 36                        Revision 2
                                                                September 1994

-------
                                   TABLE 3.
                  DFTPP KEY IONS AND ION ABUNDANCE CRITERIA8'"
       Mass
Ion Abundance Criteria
       51

       68
       70

      127

      197
      198
      199

      275

      365

      441
      442
      443
30-60% of mass 198

< 2% of mass 69
< 2% of mass 69

40-60% of mass 198

< 1% of mass 198
Base peak, 100% relative abundance
5-9% of mass 198

10-30% of mass 198

> 1% of mass 198

Present but less than mass 443
> 40% of mass 198
17-23% of mass 442
   a  See Reference 3.

   b  Alternate  tuning  criteria  may  be  used  (e.g.,  CLP,   Method  525,  or
      manufacturers'  instructions),  provided that  method performance  is  not
      adversely affected.
                                   TABLE 4.
                          CALIBRATION  CHECK COMPOUNDS
Base/Neutral Fraction
                  Acid  Fraction
Acenaphthene
1,4-Dichlorobenzene
Hexachlorobutadi ene
N-Nitrosodiphenylamine
Di-n-octyl phthalate
Fluoranthene
Benzo(a)pyrene
                  4-Chloro-3-methyl phenol
                  2,4-Dichlorophenol
                  2-Nitrophenol
                  Phenol
                  Pentachlorophenol
                  2,4,6-Trichlorophenol
                                  8270B - 37
                                            Revision  2
                                        September  1994

-------
                                   TABLE 5.
          SEMIVOLATILE  INTERNAL  STANDARDS WITH  CORRESPONDING  ANALYTES
                           ASSIGNED  FOR QUANTITATION
l,4-Dichlorobenzene-d4    Naphthalene-d8
                                Acenaphthene-d
                                                                        10
Aniline
Benzyl alcohol
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl)
                ether
2-Chlorophenol
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
Ethyl methanesulfonate
2-Fluorophenol (surr.)
Hexachloroethane
Methyl methanesulfonate
2-Methylphenol
4-Methylphenol
N-Nitrosodimethylamine
N-Nitroso-di-n-propyl-
               amine
Phenol
Phenol-de (surr.)
2-Picoline
Acetophenone
Benzoic acid
Bis(2-chloroethoxy)methane
4-Chloroaniline
4-Chloro-3-methyl phenol
2,4-Dichlorophenol
2,6-Dichlorophenol
a,a-Dimethyl-
      phenethylamine
2,4-Dimethyl phenol
Hexachlorobutadi ene
Isophorone
2-Methylnaphthalene
Naphthalene
Nitrobenzene
Nitrobenzene-d8 (surr.)
2-Nitrophenol
N-Nitrosodibutyl amine
N-Nitrosopiperidine
1,2,4-Tri chlorobenzene
Acenaphthene
Acenaphthylene
1-Chloronaphthalene
2-Chloronaphthalene
4-Chlorophenyl
  phenyl ether
Dibenzofuran
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Fluorene
2-Fluorobiphenyl
        (surr.)
Hexachlorocyclo-
      pentadiene
1-Naphthylamine
2-Naphthylamine
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
4-Nitrophenol
Pentachlorobenzene
1,2,4,5-Tetra-
   chlorobenzene
2,3,4,6-Tetra-
   chlorophenol
2,4,6-Tribromo-
   phenol  (surr.)
2,4,6-Trichloro-
   phenol
2,4,5-Trichloro-
   phenol
(surr.) = surrogate
                                  8270B  - 38
                                           Revision 2
                                       September 1994

-------
                                   TABLE 5.
                                  (Continued)
Phenanthrene-d
              10
Chrysene-d
                                    12
Perylene-d
                                                                    12
4-Aminobiphenyl
Anthracene
4-Bromophenyl phenyl
                ether
Di-n-butyl phthalate
4,6-Dinitro-2-methyl-
                phenol
Diphenylamine
Fluoranthene
Hexachlorobenzene
N-Nitrosodiphenylamine
Pentachlorophenol
Pentachloronitrobenzene
Phenacetin
Phenanthrene
Pronamide
Benzidine
Benzo(a)anthracene
Bis(Z-ethylhexyl)
       phthalate
Butyl benzyl phthalate
Chrysene
3,3'-Dichlorobenzidine
p-Dimethylaminoazobenzene
Pyrene
Terphenyl-d14  (surr.)
Benzo(b)fluor-
    anthene
Benzo(k)fluor-
    anthene
Benzo(g,h,i)-
    perylene
Benzo(a)pyrene
Dibenz(a,j)acridine
Dibenz(a,h)-
     anthracene
7,12-Dimethylbenz-
    (a)anthracene
Di-n-octyl phthalate
Indeno(l,2,3-cd)
     pyrene
3-Methylchol-
     anthrene
(surr.) = surrogate
                                  8270B - 39
                                           Revision  2
                                      September  1994

-------
       TABLE 6.
QC ACCEPTANCE CRITERIA8
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benz(a)anthracene
Benzo(b)fluoranthene
Benzo(k)fluoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Benzyl butyl phthalate
/3-BHC
5-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
2 -Chl oronaphthal ene
4-Chlorophenyl phenyl ether
Chrysene
4,4'-DDD
4, 4' -DDE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1 , 2-Di chl orobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fluoranthene
Fluorene
Heptachlor
Heptachlor epoxide
Hexachl orobenzene
Hexachlorobutadiene
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
Limit
for s
27.6
40.2
39.0
32.0
27.6
38.8
32.3
39.0
58.9
23.4
31.5
21.6
55.0
34.5
46.3
41.1
23.0
13.0
33.4
48.3
31.0
32.0
61.6
70.0
16.7
30.9
41.7
32.1
71.4
30.7
26.5
23.2
21.8
29.6
31.4
16.7
32.5
32.8
20.7
37.2
. 54.7
24.9
26.3
Range
for x
(M9/L)
60.1-132.3
53.5-126.0
7.2-152.2
43.4-118.0
41.8-133.0
42.0-140.4
25.2-145.7
31.7-148.0
D-195.0
D-139.9
41.5-130.6
D-100.0
42.9-126.0
49.2-164.7
62.8-138.6
28.9-136.8
64.9-114.4
64.5-113.5
38.4-144.7
44.1-139.9
D-134.5
19.2-119.7
D-170.6
D-199.7
8.4-111.0
48.6-112.0
16.7-153.9
37.3-105.7
8.2-212.5
44.3-119.3
D-100.0
D-100.0
47.5-126.9
68.1-136.7
18.6-131.8
D-103.5
D-188.8
42.9-121.3
71.6-108.4
D-172.2
70.9-109.4
7.8-141.5
37.8-102.2
Range
P, Ps
47-145
33-145
D-166
27-133
33-143
24-159
11-162
17-163
D-219
D-152
24-149
D-110
12-158
33-184
36-166
8-158
53-127
60-118
25-158
17-168
D-145
4-136
D-203
D-227
1-118
32-129
D-172
20-124
D-262
29-136
D-114
D-112
39-139
50-158
4-146
D-107
D-209
26-137
59-121
D-192
26.155
D-152
24-116
       8270B  -  40
    Revision 2
September 1994

-------




Compound
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitrosodi-n-propylamine
PCB-1260
Phenanthrene
Pyrene
1 , 2 , 4-Tri chl orobenzene
4-Chl oro-3-methyl phenol
2-Chlorophenol
2,4-Chlorophenol
2, 4-Dimethyl phenol
2,4-Dinitrophenol
2-Methyl -4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
TABLE 6.
(Continued)
Test
cone.
(M9/L)
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100
100


Limit
for s
(M9/L)
24.5
44.6
63.3
30.1
39.3
55.4
54.2
20.6
25.2
28.1
37.2
28.7
26.4
26.1
49.8
93.2
35.2
47.2
48.9
22.6
31.7


Range
for x
(M9/L)
55.2-100.0
D-150.9
46.6-180.2
35.6-119.6
54.3-157.6
13.6-197.9
19.3-121.0
65.2-108.7
69.6-100.0
57.3-129.2
40.8-127.9
36.2-120.4
52.5-121.7
41.8-109.0
D-172.9
53.0-100.0
45.0-166.7
13.0-106.5
38.1-151.8
16.6-100.0
52.4-129.2


Range
P> Ps
(%)
40-113
D-171
21-196
21-133
35-180
D-230
D-164
54-120
52-115
44-142
22-147
23-134
39-135
32-119
D-191
D-181
29-182
D-132
14-176
5-112
37-144
s

X


P, Ps

D

a
      Standard deviation of four recovery measurements,  in jiig/L.

      Average recovery for four recovery measurements,  in jug/L.

      Percent recovery measured.

      Detected; result must be greater than zero.

Criteria from 40 CFR  Part  136  for  Method  625.   These criteria are based
directly on the method performance  data  in Table  7.  Where necessary, the
limits for  recovery  have  been broadened to assure  applicability  of the
limits to concentrations below those used to develop Table 7.
                                  8270B - 41
                                                              Revision 2
                                                          September 1994

-------
                         TABLE 7.
METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION8
Compound
Acenaphthene
Acenaphthylene
Aldrin
Anthracene
Benz(a)anthracene
Chloroethane
Benzo(b)fluoranthene
Benzo ( k) f 1 uoranthene
Benzo(a)pyrene
Benzo(ghi)perylene
Benzyl butyl phthalate
/3-BHC
5-BHC
Bis(2-chloroethyl) ether
Bis(2-chloroethoxy)methane
Bi s ( 2-chl oroi sopropyl )
ether
Bis(2-ethylhexyl)
phthalate
4-Bromophenyl phenyl
ether
2-Chloronaphthalene
4-Chlorophenyl phenyl
ether
Chrysene
4,4'-DDD
4,4'-DDE
4,4'-DDT
Dibenzo( a, h) anthracene
Di-n-butyl phthalate
1,2-Dichlorobenzene
1 ,3-Dichl orobenzene
1,4-Dichlorobenzene
3,3'-Dichlorobenzidine
Dieldrin
Diethyl phthalate
Dimethyl phthalate
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Endosulfan sulfate
Endrin aldehyde
Fl uoranthene
Accuracy, as
recovery, x'
(M9/L)
0.96C+0.19
0.89C+0.74
0.78C+1.66
0.80C+0.68
0.88C-0.60
0.99C-1.53
0.93C-1.80
0.87C-1.56
0.90C-0.13
0.98C-0.86
0.66C-1.68
0.87C-0.94
0.29C-1.09
0.86C-1.54
1.12C-5.04
1.03C-2.31

0.84C-1.18

0.91C-1.34

0.89C+0.01
0.91C+0.53

0.93C-1.00
0.56C-0.40
0.70C-0.54
0.79C-3.28
0.88C+4.72
0.59C+0.71
0.80C+0.28
0.86C-0.70
0.73C-1.47
1.23C-12.65
0.82C-0.16
0.43C+1.00
0.20C+1.03
0.92C-4.81
1.06C-3.60
0.76C-0.79
0.39C+0.41
0.76C-3.86
0.81C+1.10
Single analyst
precision, s/
(M9/L)
0.15X-0.12
0.24x-1.06
0.27X-1.28
0.21X-0.32
O.lBx+0.93
0.14X-0.13
0.22X+0.43
0.19X+1.03
0.22X+0.48
0.29X+2.40
O.lSx+0.94
0.20X-0.58
0.34X+0.86
0.35X-0.99
0.16X+1.34
0.24X+0.28

0.26X+0.73

0.13X+0.66

0.07X+0.52
0.20X-0.94

0.28X+0.13
0.29X-0.32
0.26X-1.17
0.42X+0.19
0.30X+8.51
0.13X+1.16
0.20X+0.47
0.25X+0.68
0.24x+0.23
0.28X+7.33
0.20X-0.16
0.28X+1.44
0.54x+0.19
0.12X+1.06
0.14X+1.26
0.21X+1.19
0.12X+2.47
O.lSx+3.91
0.22X-0.73
Overall
precision,
S' (M9/L)
0.21x-0.67
0.26x-0.54
0.43X+1.13
0.27x-0.64
0.26x-0.21
O.Ux-0.28
0.29x+0.96
0.35x+0.40
0.32X+1.35
O.Blx-0.44
0.53x+0.92
0.30x+1.94
0.93x-0.17
0.35x+0.10
0.26x+2.01
0.25X+1.04

0.36X+0.67

0.16x+0.66

0.13x+0.34
0.30X-0.46

0.33X-0.09
0.66X-0.96
0.39x-1.04
0.65X-0.58
0.59X+0.25
0.39x+0.60
0.24x+0.39
0.41X+0.11
0.29x+0.36
0.47x+3.45
0.26x-0.07
0.52X+0.22
l.OBx-0.92
0.21x+1.50
0.19x+0.35
0.37x+1.19
0.63x-1.03
0.73x-0.62
0.28x-0.60
                        8270B  - 42
    Revision 2
September 1994

-------




Compound
Fluorene
Heptachlor
Heptachlor epoxide
Hexachlorobenzene
Hexachlorobutadiene
Hexachloroethane
Indeno(l,2,3-cd)pyrene
Isophorone
Naphthalene
Nitrobenzene
N-Nitrosodi -n-propylamine
PCB-1260
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
4-Chloro-3 -methyl phenol
2-Chlorophenol
2,4-Dichlorophenol
2,4-Dimethylphenol
2,4-Dinitrophenol
2-Methyl-4,6-dinitrophenol
2-Nitrophenol
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
TABLE 7.
(Continued)
Accuracy, as
recovery, x'
(M9/L)
0.90C-0.00
0.87C-2.97
0.92C-1.87
0.74C+0.66
0.71C-1.01
0.73C-0.83
0.78C-3.10
1.12C+1.41
0.76C+1.58
1.09C-3.05
1.12C-6.22
0.81C-10.86
0.87C+0.06
0.84C-0.16
0.94C-0.79
0.84C+0.35
0.78C+0.29
0.87C-0.13
0.71C+4.41
0.81C-18.04
1.04C-28.04
0.07C-1.15
0.61C-1.22
0.93C+1.99
0.43C+1.26
0.91C-0.18


Single analyst
precision, s/
(M9/L)
0.12X+0.26
0.24X-0.56
0.33X-0.46
O.lSx-0.10
0.19X+0.92
0.17X+0.67
0.29x+1.46
0.27x+0.77
O.Zlx-0.41
0.19x+0.92
0.27x+0.68
0.35X+3.61
0.12X+0.57
0.16X+0.06
O.lSx+0.85
0.23x+0.75
O.lSx+1.46
0.15X+1.25
0.16X+1.21
0.38X+2.36
O.lOx+42.29
0.16X+1.94
0.38X+2.57
0.24X+3.03
0.26X+0.73
0.16X+2.22


Overall
precision,
S' (M9/L)
0.13X+0.61
O.SOx-0.23
0.28X+0.64
0.43X-0.52
0.26X+0.49
0.17X+0.80
O.SOx-0.44
0.33X+0.26
0.30X-0.68
0.27x+0.21
0.44X+0.47
0.43X+1.82
O.lSx+0.25
O.lBx+0.31
0.21X+0.39
0.29X+1.31
0.28X+0.97
0.21X+1.28
0.22X+1.31
0.42X+26.29
0.26X+23.10
0.27X+2.60
0.44X+3.24
0.30X+4.33
0.35X+0.58
0.22X+1.81
X'
S'


C

x
Expected  recovery  for  one  or  more  measurements  of  a  sample
containing a concentration of C,  in M9/L-

Expected single  analyst  s_tandard  deviation of measurements at  an
average concentration of x,  in M9/L.

Expected interlaboratory standard deviation of measurements at  an
average concentration found of x,  in M9/L-

True value for the concentration,  in M9/L-

Average recovery  found for measurements  of  samples containing a
concentration of C, in M9/L.
                                  8270B - 43
                                                       Revision 2
                                                   September  1994

-------
                                   TABLE 8.
      SURROGATE  SPIKE  RECOVERY  LIMITS  FOR  WATER  AND  SOIL/SEDIMENT  SAMPLES

Surrogate Compound
Nitrobenzene-d5
2-Fluorobiphenyl
Terphenyl-d14
Phenol -d6
2-Fluorophenol
2,4,6-Tribromophenol
Low/High
Water
35-114
43-116
33-141
10-94
21-100
10-123
Low/High
Soil/Sediment
23-120
30-115
18-137
24-113
25-121
19-122
                                   TABLE 9.
              EXTRACTION  EFFICIENCY  AND  AQUEOUS  STABILITY  RESULTS
COMPOUND
PERCENT RECOVERY
   ON DAY 0
AVG.      RSD
PERCENT RECOVERY
  ON DAY 7
AVG.    RSD
3-Amino-9-ethylcarbazole
4-Chloro-l,2-phenylenediamine
4-Chloro-l,3-phenylenediamine
1 , 2-Di bromo-3-chl oropropane
2-sec-Butyl-4,6-diriitrophenol
Ethyl parathion
4 ,4' -Methyl enebis(N,N-di methyl aniline)
2-Methyl-5-nitroaniline
2-Methylpyridine
Tetraethyl dithiopyrophosphate
80
91
84
97
99
100
108
99
80
92
8
1
3
2
3
2
4
10
4
7
73
108
70
98
97
103
90
93
83
70
3
4
3
5
6
4
4
4
4
1
Data from Reference 8.
                                  8270B - 44
                            Revision  2
                        September  1994

-------
                              TABLE 10.
AVERAGE PERCENT RECOVERIES AND PERCENT RSDs FOR THE TARGET COMPOUNDS
  FROM  SPIKED CLAY  SOIL AND TOPSOIL BY  AUTOMATED  SOXHLET EXTRACTION
                     WITH  HEXANE-ACETONE  (1:1)'
                                   Clay Soil
 Topsoil
Compound name
1,3-Dichlorobenzene
1,2-Dichlorobenzene
Nitrobenzene
Benzal chloride
Benzotrichloride
4-Chloro-2-nitrotoluene
Hexachlorocyclopentadiene
2,4-Dichloronitrobenzene
3 , 4-Di chl oron i trobenzene
Pentachl orobenzene
2, 3, 4, 5-Tetrachl oron i trobenzene
Benefin
alpha-BHC
Hexachl orobenzene
delta-BHC
Heptachlor
Aldrin
Isopropalin
Heptachlor epoxide
trans-Chlordane
Endosulfan I
Dieldrin
2,5-Dichlorophenyl-
4-nitrophenyl ether
Endrin
Endosulfan II
p,p'-DDT
2,3,6-Trichlorophenyl-
4'-nitrophenyl ether
2 , 3 , 4-Tri chl orophenyl -
4'-nitrophenyl ether
Mi rex
Average
percent
recovery
0
0
0
0
0
0
4.1
35.2
34.9
13.7
55.9
62.6
58.2
26.9
95.8
46.9
97.7
102
90.4
90.1
96.3
129
110

102
104
134
110

112

104
Percent
RSD
..
--
--
--
--
--
15
7.6
15
7.3
6.7
4.8
7.3
13
4.6
9.2
12
4.3
4.4
4.5
4.4
4.7
4.1

4.5
4.1
2.1
4.8

4.4

5.3
Average
percent
recovery
0
0
0
0
0
0
7.8
21.2
20.4
14.8
50.4
62.7
54.8
25.1
99.2
49.1
102
105
93.6
95.0
101
104
112

106
105
111
110

112

108
Percent
RSD
..
--
--
--
--
--
23
15
11
13
6.0
2.9
4.8
5.7
1.3
6.3
7.4
2.3
2.4
2.3
2.2
1.9
2.1

3.7
0.4
2.0
2.8

3.3

2.2
 The  operating conditions  for  the Soxtec  apparatus  were as  follows:
 immersion time 45 min; extraction time 45 min; the sample size was 10 g;
 the spiking concentration was 500 ng/g, except  for the surrogate compounds
 at 1000 ng/g, compounds 23, 27,  and 28  at  1500  ng/g,  compound 3 at 2000
 ng/g> and compounds 1 and 2 at 5000 ng/g.
                             8270B - 45
    Revision 2
September 1994

-------
                           TABLE  11.
SINGLE LABORATORY ACCURACY AND PRECISION DATA FOR THE EXTRACTION
          OF SEMIVOLATILE ORGANICS FROM SPIKED CLAY BY
                METHOD 3541 (AUTOMATED SOXHLET)"
Compound name
Phenol
Bis(2-chloroethy1)ether
2-Chlorophenol
Benzyl alcohol
2-Methyl phenol
Bis(2-chloroisopropyl )ether
4-Methyl phenol
N-Nitroso-di-n-propylamine
Nitrobenzene
Isophorone
2-Nitrophenol
2,4-Dimethylphenol
Benzoic acid
Bis(2-chloroethoxy)methane
2,4-Dichlorophenol
1 , 2 , 4-Tr i chl orobenzene
Naphthalene
4-Chloroaniline
4-Chl oro-3-methyl phenol
2-Methyl naphthal ene
Hexachl orocycl opentadi ene
2,4, 6 -Tr i chl orophenol
2,4,5-Trichlorophenol
2-Chloronaphthalene
2-Nitroaniline
Dimethyl phthalate
Acenaphthylene
3-Nitroaniline
Acenaphthene
2,4-Dinitrophenol
4-Nitrophenol
Dibenzofuran
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Diethyl phthalate
4-Chlorophenyl-phenyl ether
Fluorene
4-Nitroaniline
4 , 6 -Di ni tro -2 -methyl phenol
N-Nitrosodiphenylamine
4-Bromophenyl-phenyl ether
Average
percent
recovery
47.8
25.4
42.7
55.9
17.6
15.0
23.4
41.4
28.2
56.1
36.0
50.1
40.6
44.1
55.6
18.1
26.2
55.7
65.1
47.0
19.3
70.2
26.8
61.2
73.8
74.6
71.6
77.6
79.2
91.9
62.9
82.1
84.2
68.3
74.9
67.2
82.1
79.0
63.4
77.0
62.4
Percent
RSD
5.6
13
4.3
7.2
6.6
15
6.7
6.2
7.7
4.2
6.5
5.7
7.7
3.0
4.6
31
15
12
5.1
8.6
19
6.3
2.9
6.0
6.0
5.2
5.7
5.3
4.0
8.9
16
5.9
5.4
5.8
5.4
3.2
3.4
7.9
6.8
3.4
3.0
                           8270B - 46
    Revision 2
September 1994

-------
                             Table  11.  (Continued)
                                                Average
                                                percent           Percent
Compound name                                   recovery            RSD


Hexachlorobenzene                                 72.6              3.7
Pentachlorophenol                                 62.7              6.1
Phenanthrene                                      83.9              5.4
Anthracene                                        96.3              3.9
Di-n-butyl phthalate                              78.3             40
Fluoranthene                                      87.7              6.9
Pyrene                                           102                0.8
Butyl benzyl phthalate                            66.3              5.2
3,3'-Dichlorobenzidine                            25.2             11
Benzo(a)anthracene                                73.4              3.8
Bis(2-ethylhexyl) phthalate                       77.2              4.8
Chrysene                                          76.2              4.4
Di-n-octyl phthalate                              83.1              4.8
Benzo(b)fluoranthene                              82.7              5.0
Benzo(k)fluoranthene                              71.7              4.1
Benzo(a)pyrene                                    71.7              4.1
Indeno(l,2,3-cd)pyrene                            72.2              4.3
Dibenzo(a,h)anthracene                            66.7              6.3
Benzo(g,h,i)perylene                              63.9              8.0
1,2-Dichlorobenzene                                0
1,3-Dichlorobenzene                                0
1,4-Dichlorobenzene                                0
Hexachloroethane                                   0
Hexachlorobutadiene                                0
a     Number  of  determinations was three.   The operating conditions  for  the
      Soxtec apparatus were as follows:   immersion time 45 min; extraction time
      45 min; the sample  size  was  10  g  clay  soil; the  spike  concentration  was
      6 mg/kg per compound.  The sample  was allowed to equilibrate 1 hour after
      spiking.

Data taken from Reference 9.
                                  8270B - 47                        Revision  2
                                                                September 1994

-------
                             FIGURE 1.
GAS CHROMATOGRAM OF BASE/NEUTRAL AND ACID CALIBRATION  STANDARD
                          : 5lbH*C«w786 kl
                      CnLl: 51BHS668786 13
SCANS  2t* TO 270e
KIC
06/u7'86  8:26:08
StWPLE: BASE ACID STD, 2UL/;.uHC,. UL
CODS.:
R»«CE: C   1.2780 Lfl££L: N 6. 4.9  CKJHN: M 0- t.d J 6  bkS£: U 2d-   3
                                                                JUL
                                                   33:20
                          41:46
                            8270B - 48
                       Revision 2
                   September 1994

-------
                                  METHOD  8270B
SEMIVOLATILE ORGANIC COMPOUNDS  BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY
                     (GC/MS):   CAPILLARY COLUMN TECHNIQUE
    7.1 Prepare sample
    using Method 3540,
      3541, or 3550.
7.1 Prepare sample
using Method 3510
    or 3520.
                                  7.1 Prepare sample
                                  using Method 3540,
                                 3541, 3550, or 3580.
                                     7.2 Cleanup
                                      extract.
                                   7.3 Set GC/MS
                                 operating conditions;
                                   perform initial
                                     calibration.
                                   7.4 Perform daily
                                 calibration with SPCCs
                                   and CCCs prior to
                                  analysis of samples.
                                   8270B  - 49
            Revision 2
       September 1994

-------
                                    METHOD 8270B
                                     (Continued)
                            7.5.1  Screen extract
                           on GC/FID or GC/PID to
                            eliminate samples that
                            are too concentrated.
                            7.5.3 Analyze extract
                              by GC/MS, using
                            appropriate fused-silica
                              capillary column.
7.S.4 Dilute
  Extract.
    7.5.4
Does response
 exceed initial
  calibration
   curve?
^
No
r
7.6.1 Identify
analyte by comparing
the sample and standard
mass spectra.
^1
r
7.6.2 Calculate
concentration of each
individual analyte;
report results.
>
r
                                C   Stop   "\
                                     8270B  -  50
                                                 Revision  2
                                            September 1994

-------
                      ERRATA FOR METHOD 8280


In Section 1.1, delete the following text:
      "reactor residues" with no replacement.

In Section 1.5, replace the following text:
      "the analyst must take necessary  precautions to prevent exposure to
himself, or to others, of"
with:
      "the  analyst must  take  necessary  precautions  to  prevent  human
exposure from" and
delete the following text:
      "to be  reviewed  and approved by EPA's  Dioxin  Task Force (Contact
Conrad Kleveno, WH 548A,  U.S.  EPA, 401 M  Street  S.W.,  Washington, D.C.
20450)."

In Section 6.3, replace the following text:
      "x = measured as in Figure 2"
with:
      "x =  height of the valley between  2,3,7,8-TCDD  and 1,2,3,4-TCDD,
using the column performance check mixture."

In Section 6.9.2, replace "a 2-hr period" with "a 12 hr period".
In Section 7.4, replace "24" with "20".
                          8280 ERRATA -  1                       July 1992

-------
                                 METHOD 8280

              THE ANALYSIS OF POLYCHLORINATED DIBENZO-P-DIOXINS
                      AND POLYCHLORINATED DIBENZOFURANS
1.0  SCOPE AND APPLICATION

     1.1  This method is appropriate for  the determination of tetra-,  penta-,
hexa-,  hepta-,  and  octachlorinated  dibenzo-p-dioxins  (PCDD's)  and dibenzo-
furans   (PCDF's)  in  chemical  wastes  including  still   bottoms,  fuel  oils,
sludges, fly ash, reactor residues, soil and water.

     1.2  The sensitivity  of  this  method  is  dependent  upon  the level  of
interferents within a given matrix.  Proposed quantification levels for target
analytes were 2 ppb in soil samples, up to 10 ppb in other solid wastes and
10 ppt  in water.  Actual values  have  been shown to vary by homologous series
and, to a lesser degree, by individual  isomer.  The total detection limit for
each CDD/CDF homologous  series  is  determined  by  multiplying the detection
limit of a given isomer within that series by the number of peaks which can  be
resolved under the gas chromatographic conditions.

     1.3  Certain   2,3,7,8-substituted   congeners   are   used   to  provide
calibration and method  recovery  information.    Proper  column selection and
access  to reference isomer  standards,  may  in  certain cases, provide isomer
specific data.    Special  instructions  are  included  which measure 2,3,7,8-
substituted congeners.

     1.4  This method is recommended for use only by analysts experienced with
residue analysis and skilled in mass spectral analytical  techniques.

     1.5  Because of the extreme toxicity of these compounds, the analyst must
take necessary precautions to prevent  exposure  to  himself, or to others,  of
materials known or believed to  contain  PCDD's or PCDF's.  Typical infectious
waste incinerators  are  probably  not  satisfactory  devices  for disposal  of
materials highly contaminated with PCDD's or PCDF's.  A laboratory planning to
use these compounds should prepare a disposal plan to be reviewed and approved
by  EPA's Dioxin Task Force (Contact  Conrad  Kleveno, WH-548A, U.S. EPA, 401 M
Street S.W., Washington,  D.C.  20450).    Additional  safety instructions are
outlined in Appendix B.


2.0  SUMMARY OF THE METHOD

     2.1  This procedure  uses  a  matrix-specific extraction, analyte-specific
cleanup,  and   high-resolution   capillary   column   gas  chromatography/low
resolution mass spectrometry  (HRGC/LRMS) techniques.

     2.2  If   interferents  are  encountered,  the  method  provides  selected
cleanup procedures to aid  the analyst  in their elimination.   The  analysis flow
chart is shown  in  Figure  1.
                                  8280 -  1
                                                         Revision      0
                                                         Date  September 1986

-------
                      Complex
                       Waste
                      Sample
                            (1)   Add Internal  Standards:   13C12-PCDD's
                                 and 13C12-PCDF's.

                            (2)   Perform matrix-specific  extraction.
                      Sample
                      Extract
                            (1)   Wash with 20% KOH
                            (2)   Wash with 5% NaCl
                            (3)   Wash with cone. H2S04
                            (4)   Wash with 5% NaCl
                            (5)   Dry extract
                            (6)   Solvent exchange
                            (7)   Alumina column
                  60% CH2Cl2/hexane
                     Fraction
                            (1)  Concentrate eluate
                            (2)  Perform carbon column cleanup
                            (3)  Add recovery standard(s)-13C12-l,2,3,4-TCDD
                  Analyze by GC/MS
Figure 1.  Method 8280 flow chart for sample extraction and cleanup as
 used for the analysis of PCDD's and PCDF's 1n complex waste samples.
                           8280 - 2
                                                  Revision      0
                                                  Date   September 1986

-------
3.0  INTERFERENCES

     3.1  Solvents, reagents, glassware,  and  other sample processing hardware
may  yield  discrete  artifacts  and/or  elevated  baselines  which  may cause
misinterpretation of chromatographic data.    All  of  these materials must be
demonstrated to be free from interferents  under the conditions of analysis by
running laboratory method blanks.

     3.2  The use of  high  purity  reagents  and  solvents  helps to minimize
interference problems.  Purification of  solvents by distillation in all glass
systems may be required.

     3.3  Interferents co-extracted  from  the  sample  will vary considerably
from source to source,  depending  upon  the industrial process being sampled.
PCDD's and PCDF's  are  often  associated  with  other interfering chlorinated
compounds such as  PCB's and polychlorinated diphenyl ethers which may be found
at concentrations  several orders of magnitude higher than that of the analytes
of interest.   Retention  times  of  target  analytes  must  be verified using
reference standards.   These  values  must  correspond  to  the retention time
windows established  in  Section  6-3.    While  certain cleanup techniques are
provided as part of  this method, unique  samples may require additional  cleanup
techniques to achieve  the  method  detection  limit   (Section 11.6) stated in
Table 8.

     3.4  High resolution capillary columns are  used  to resolve as many PCDD
and  PCDF isomers as  possible;  however,   no  single column  is  known to  resolve
all  of the isomers.

     3.5  Aqueous  samples cannot  be  aliquoted  from  sample  containers.  The
entire sample must be  used  and the sample container washed/rinsed out with the
extracting solvent.


4.0  APPARATUS AND MATERIALS

     4.1  Sampling equipment  for discrete or composite sampling;

          4.1.1  Grab  sample  bottle—amber  glass,  1-liter or 1-quart  volume.
     French or Boston  Round design is recommended.  The container must  be acid
     washed and  solvent rinsed before use to minimize  interferences.

          4.1.2  Bottle caps--threaded to screw  onto the  sample bottles.  Caps
     must be  lined with Teflon.  Solvent washed  foil,  used  with the shiny side
     toward the  sample, may be   substituted  for  Teflon  if the sample is not
     corrosive.  Apply tape around cap to completely seal cap  to bottom.

          4.1.3  Compositing   equipment—automatic    or    manual  compositing
     system.  No tygon or   rubber  tubing  may  be  used,  and the system must
     incorporate glass sample containers for  the collection of a minimum of
     250 ml.  Sample containers  must be  kept refrigerated after  sampling.

     4.2  Water  bath—heated,   with   concentric   ring    cover,  capable  of
temperature control  (+2*C).  The bath  should be  used in a hood.

                                  8280 - 3
                                                          Revision       0
                                                          Date  September 1986

-------
4.3  Gas chromatograph/mass spectrometer data system;

     4.3.1  Gas chromatograph:  An  analytical system with a temperature-
programmable gas  chromatograph  and  all  required accessories including
syringes, analytical columns, and gases.

     4.3.2  Fused silica capillary  columns  are  required.   As shown in
Table 1, three columns  were  evaluated  using a column performance check
mixture   containing   1,2,3,4-TCDD,   2,3,7,8-TCDD,   1,2,3,4,7   PeCDD,
1,2,3,4,7,8-HxCDD, 1,2,3,4,6,7,8-HpCDD, OCDD, and 2,3,7,8-TCDF.

The columns include the  following:     (a) 50-m CP-Sil-88 programmed 60*-
190* at 20*/minute, then 190*-240* at 5*/minute; (b) DB-5  (30-m x 0.25-mm
I.D.; 0.25-um film thickness) programmed  170* for 10 minutes, then 170*-
320* at  8*/minute,  hold  at  320*C  for  20  minutes;   (c) 30-m SP-2250
programmed 70*-320* at  10*/minute.     Column/conditions  (a) provide good
separation of 2,3,7,8-TCDD from the other TCDD's at  the expense of longer
retention times  for higher homologs.    Column/conditions  (b) and  (c) can
also  provide  acceptable  separation   of  2,3,7,8-TCDD.    Resolution of
2,3,7,8-TCDD from the other  TCDD's  is  better  on column  (c), but column
 (b)  is   more  rugged,  and   may   provide  better  separation  from  certain
classes  of  interferents.  Data presented  in  Figure 2 and  Tables 1  to 8 of
this  Method     were  obtained  using   a  DB-5  column  with  temperature
programming described in  (b)  above.    However, any capillary  column which
provides separation  of  2,3,7,8-TCDD   from  all  other TCDD   isomers
equivalent  to that  specified in Section  6.3 may be  used;  this  separation
must be  demonstrated  and  documented   using   the performance  test  mixture
described in  Paragraph 6.3.

     4.3.3  Mass spectrometer:  A low  resolution instrument is  specified,
 utilizing 70   volts  (nominal)  electron   energy   in  the electron impact
 ionization  mode.  The system  must be capable  of  selected ion  monitoring
 (SIM)  for at  least  11 ions  simultaneously,   with a  cycle  time of  1 sec or
 less.   Minimum  integration  time  for SIM  is 50  ms per  m/z.  The use of
 systems  not capable of monitoring 11 ions simultaneously  will  require the
 analyst  to  make multiple  injections.

     4.3.4  GC/MS  Interface:    Any   GC-to-MS   interface  that  gives an
 acceptable  calibration   response   for   each   analyte  of   interest at the
 concentration  required   and  achieves  the   required  tuning performance
 criteria (see Paragraphs   6.1.-6.3)  may  be used.    GC-to-MS  interfaces
 constructed of all  glass   or  glass-lined materials are  required.  Glass
 can be deactivated by  silanizing with dichlorodimethyl si lane.   Inserting
 a fused silica  column directly  into   the MS source is recommended;  care
 must be taken not to expose the  end of the column  to the  electron beam.

     4.3.5  Data system:   A  computer   system  must  be interfaced to the
 mass spectrometer.   The  system  must allow for  the  continuous acquisition
 and storage on machine-readable  media  of all data  obtained throughout the
 duration of the chromatographic  program.  The computer must have  software
 that can search any GC/MS data file  for  ions  of  a specific mass and can
 plot such ion abundances  versus  time   or  scan  number.  This  type of plot
                              8280 - 4
                                                     Revision      0
                                                     Date  September 1986

-------
     Is defined as an Selected Ion Current Profile (SICP).   Software must also
     be able to integrate the  abundance,   in any SICP,  between specified time
     or scan number limits.

     4.4  Pipets-Disposable,  Pasteur,  150-mm   long   x  5-mrn  I.D.  (Fisher
Scientific Company, No. 13-678-6A, or equivalent).

          4.4.1  Pipet,  disposable,  serological  10-mL  (American Scientific
     Products No.  P4644-10,  or  equivalent)  for  preparation  of the carbon
     column specified in Paragraph 4.19.

     4.5  Amber glass bottle (500-mL, Teflon-lined screw-cap).

     4.6  Reacti-vial 2-mL,  amber  glass  (Pierce  Chemical  Company).  These
should be silanized prior to use.

     4.7  500-mL Erlenmeyer flask (American Scientific Products Cat. No. f4295
500fO) fitted with Teflon stoppers  (ASP No. S9058-8, or equivalent).

     4.8  Wrist Action Shaker  (VWR No. 57040-049, or equivalent).

     4.9  125-mL   and  2-L  Separatory  Funnels   (Fisher  Scientific  Company,
No.  10-437-5b, or  equivalent).

     4.10   500-mL  Kuderna-Danish  fitted with a 10-mL concentrator tube and
3-ball Snyder column  (Ace Glass No.  6707-02, 6707-12, 6575-02, or equivalent).

     4.11   Teflon  boiling chips  (Berghof/American Inc., Main  St., Raymond, New
Hampshire 03077, No.  15021-450,   or  equivalent).    Wash with hexane prior to
use.

     4.12   300-mm  x  10.5-mm glass   chromatographic  column fitted  with  Teflon
stopcock.

     4.13   15-mL  conical   concentrator    tubes    (Kontes   No.  K-288250,  or
equivalent).

     4.14   Adaptors  for  concentrator  tubes   (14/20  to  19/22)  (Ace Glass No.
9092-20, or equivalent).

     4.15   Nitrogen  blowdown   apparatus   (N-Evap  (reg.  trademark)  Analytical
Evaporator   Model   111,    Organomation   Associates   Inc.,   Northborough,
Massachusetts  or  equivalent).    Teflon   tubing connection   to   trap  and gas
regulator  is required.

     4.16   Microflex conical  vials  2.0-mL (Kontes K-749000, or equivalent).

     4.17   Filter paper  (Whatman  No. 54,   or equivalent).   Glass fiber  filters
or glass wool  plugs  are  also  recommended.

     4.18   Solvent reservoir  (125-mL)  Kontes;     (special  order item)  12.5-cm
diameter, compatible with  gravity carbon  column.


                                  8280 -  5
                                                         Revision      0
                                                          Date   September  1986

-------
     4.19  Carbon column (gravity  flow);    Prepare carbon/silica gel packing
material by  mixing5percent  (byweight)  active  carbon  AX-21 (Anderson
Development Co., Adrain,  Michigan),  pre-washed  with  methanol  and dried Iji
vacuo at 110'C and 95 percent (by  weight)  Silica gel (Type 60, EM reagent 70
to 230 mesh, CMS No. 393-066)  followed  by  activation of the mixture at 130*
for 6 hr.  Prepare a  10-mL  disposable  serologlcal pipet by cutting off each
end to achieve a  4-1n.  column.    F1re  polish  both ends; flare If desired.
Insert a glass-wool plug at one end and pack with 1 g of the carbon/siHca gel
mixture.  Cap the packing with a glass-wool plug.  (Attach reservoir to column
for addition of solvents).

     Option:  Carbon column  (HPLC):  A  silanized glass HPLC column (10 mm x 7
cm), or equivalent, which  contains  1  g  of  a  packing prepared by mixing 5
percent (by weight) active  carbon  AX-21,   (Anderson Development Co., Adrian,
Michigan), washed with methanol and  dried   iji  vacuo at 110'C, and 95 percent
(by weight)  10  urn  silica  (Spherisorb  S10W  from  Phase Separations, Inc.,
Norwalk, Connecticut).  The mixture must  then be stirred and sieved through a
38-um screen (U.S. Sieve  Designation  400-mesh, American Scientific Products,
No. S1212-400, or equivalent) to remove any  clumps.1

     4.20  HPLC pump with loop  valve  (1.0  ml)  injector  to  be used in the
optional carbon column cleanup procedure.

     4.21  Dean-Stark trap,  5-  or  10-mL  with  T  joints, (Fisher Scientific
Company, No. 09-146-5, or equivalent) condenser and 125-mL  flask.

     4.22  Continuous liquid-liquid extractor  (Hershberg-Wolfe  type,  Lab Glass
No. LG-6915; or equivalent.).

     4.23  Roto-evaporator,  R-110.   Buchi/Brinkman   - American Scientific No.
E5045-10;  or equivalent.


5.0  REAGENTS

     5.1   Potassium hydroxide  (ASC):   20 percent  (w/v)  in  distilled water.

     5.2   Sulfuric acid (ACS),  concentrated.

     5.3   Methylene  chloride,   hexane,   benzene,   petroleum  ether,  methanol,
tridecane, isooctane,  toluene,   cyclohexane.    Distilled   in glass  or highest
available purity.

      5.4  Prepare stock standards  in   a  glovebox  from  concentrates or neat
materials.  The stock solutions (50  ppm)  are  stored in  the dark at 4*C,  and
checked frequently for signs  of  degradation  or evaporation,  especially just
prior  to the preparation of working standards.
 1    The carbon column preparation and  use  is adapted from W.  A. Korfmacher,
  L. G. Rushing, D. M. Nestorick, H. C. Thompson, Jr.,  R. K.  Mltchum, and J. R.
  Kominsky,  Journal  of  High  Resolution  Chromatography  and  Chromatography
  Communications, 8, 12-19 (1985).

                                   8280 - 6
                                                          Revision      0
                                                          Date  September 1986

-------
     5.5  Alumina, neutral, Super 1, Woelm,   80/200  mesh.    Store in a sealed
container at room temperature in a desiccator over self-indicating silica gel.

     5.6  Prepurified nitrogen gas.

     5.7  Anhydrous sodium  sulfate  (reagent  grade):    Extracted  by manual
shaking with several portions of hexane and dried at 100*C.

     5.8  Sodium chloride - (analytical reagent), 5 percent (w/v) in distilled
water.
6.0  CALIBRATION

     6.1  Two types of calibration procedures are required.  One type, initial
calibration, is required  before  any  samples  are  analyzed  and is required
intermittently throughout sample analyses  as  dictated  by results of routine
calibration procedures described below.   The other type, routine calibration,
consists  of  analyzing   the   column   performance   check  solution  and  a
concentration calibration solution of 500  ng/mL  (Paragraph 6.2).  No samples
are to be analyzed until acceptable calibration as described in Paragraphs 6.3
and 6.6 is demonstrated and documented.

     6.2  Initial calibration;

          6.2.1     Prepare multi-level calibration  standards2 keeping one of
the recovery standards and the  internal standard at fixed concentrations (500
ng/mL).    Additional   internal   standards   (13Ci2-OCDD  1,000  ng/mL)  are
recommended when quantification of  the  hepta-  and octa-isomers is required.
The use of separate  internal  standards  for  the PCDF's is also recommended.
Each calibration standard should contain the following compounds:

2,3,7,8-TCDD,
1,2,3,7,8-PeCDD      or any available   2,3,7,8,X-PeCDD  isomer,
1,2,3,4,7,8-HxCDD    or any available   2,3,7,8,X,Y-HxCDD isomer,
1,2,3,4,6,7,8-HpCDD  or any available   2,3,7,8,X,Y,Z-HpCDD isomer,

2,3,7,8-TCDF
l,2,3,7,8,PeCDF      or any available   2,3,7,8,X-PeCDF  isomer,
1,2,3,4,7,8-HxCDF    or any available   2,3,7,8,X,YlHxCDF isomer,
1,2,3,4,6,7,8-HpCDF  or any available   2,3,7,8,X,Y,Z-HpCDF isomer,

     OCDD, OCDF, 13Ci2-2,3,7,8-TCDD, i3Ci2-l,2,3,4-TCDD  and i3Ci2-OCDD.
 2     13Cj2-labeled analytes  are  available   from  Cambridge  Isotope  Laboratory,
 Woburn,  Massachusetts.   Proper quantification  requires  the  use  of a  specific
  labeled  isomer for each  congener to   be  determined.   When labeled PCDD's and
  PCDF's of each homolog are   available,   their  use will be  required consistent
 with the technique of isotopic dilution.
                                   8280 - 7
                                                          Revision      0
                                                          Date   September  1986

-------
Recommended concentration levels for  standard  analytes  are 200,  500,  1,000,
2,000, and 5,000 ng/mL.  These values  may be adjusted in order to  insure that
the  analyte  concentration  falls  within  the  calibration  range.    Two uL
injections of calibration  standards  should  be  made.    However, some GC/MS
instruments may require the use of  a 1-uL injection volume; if this injection
volume 1s used then  all  injections  of  standards, sample extracts and blank
extracts must also be made at  this injection volume.  Calculation  of relative
response factors is described in Paragraph 11.1.2.  Standards must  be analyzed
using the  same  solvent  as  used  1n  the  final  sample  extract.   A wider
calibration range is  useful  for  higher  level  samples  provided  it can be
described within  the   linear  range  of  the  method,  and the identification
criteria defined in Paragraph 10.4 are  met.   All standards must be stored in
an isolated  refrigerator  at  4*C  and  protected  from  light.   Calibration
standard solutions must be replaced routinely after six months.

      6.3  Establish operating parameters for  the GC/MS system; the instrument
should be tuned to meet  the  isotopic  ratio  criteria  listed in Table 3 for
PCDD's and PCDF's.   Once  tuning  and  mass  calibration procedures have been
completed, a column performance  check  mixture3 containing the isomers listed
below should be injected into the GC/MS system:

TCDD      1,3,6,8; 1,2,8,9; 2,3,7,8; 1,2,3,4; 1,2,3,7; 1,2,3,9
PeCDD    1,2,4,6,8; 1,2,3,8,9
HxCDD    1,2,3,4,6,9;  1,2,3,4,6,7
HpCDD    1,2,3,4,6,7,8; 1,2,3,4,6,7,9
OCDD      1,2,3,4,6,7,8,9

TCDF      1,3,6,8; 1,2,8,9
PeCDF    1,3,4,6,8; 1,2,3,8,9
HxCDF    1,2,3,4,6,8;  1,2,3,4,8',9
HpCDF    1,2,3,4,6,7,8; 1,2,3,4,7,8,9
OCDF      1,2,3,4,6,7,8,9

      Because of the known  overlap  between the  late-eluting tetra-isomers and
the early-eluting penta-isomers  under  certain  column  conditions, it may be
necessary to perform two   injections  to  define the  TCDD/TCDF  and PeCDD/PeCDF
elution windows, respectively.   Use  of  this   performance  check  mixture  will
enable the  following parameters  to be   checked:   (a)  the retention windows for
each  of the  homologues,  (b)  the  GC   resolution of  2,3,7,8-TCDD  and 1,2,3,4-
TCDD, and  (c)  the  relative ion  abundance  criteria listed for PCDD's and  PCDF's
in Table  3.  GC column performance   should  be checked dally  for resolution and
peak  shape  using this  check mixture.

      The  chromatographic peak separation  between 2,3,7,8-TCDD  and  1,2,3,4-TCDD
must  be  resolved with  a valley  of <£25  percent,  where

           Valley  Percent = (x/y) (100)

      x  =  measured  as  in Figure  2
      y  =  the peak  height of 2,3,7,8-TCDD
 3    Performance check mixtures  are  available  from Brehm Laboratory,  Wright
  State University, Dayton, Ohio.

                                   8280 - 8
                                                          Revision      0
                                                          Date  September 1986

-------
     It 1s the  responsibility  of  the  laboratory  to  verify the conditions
suitable for maximum resolution of  2,3,7,8-TCDD  from all  other TCDD isomers.
The peak representing 2,3,7,8-TCDD should be labeled and identified as such on
all chromatograms.

     6.4  Acceptable  SIM  sensitivity  is  verified  by  achieving  a minimum
signal-to-noise ratio of 50:1  for  the  m/z  320 ion of 2,3,7,8-TCDD obtained
from injection of the 200 ng/mL calibration standard.

     6.5  From  injections  of  the  5  calibration  standards,  calculate the
relative response factors  (RRF's)  of  analytes  vs. the appropriate internal
standards, as described in  Paragraph  11.1.2.   Relative response factors for
the hepta- and octa-chlorinated CDD's and CDF's are to be calculated using the
corresponding 13Cj2-octachlorinated standards.

     6.6  For each analyte calculate the  mean relative response factor (RRF),
the standard  deviation,  and  the  percent  relative  standard deviation from
triplicate determinations of   relative  response  factors for each calibration
standard solution.

     6.7  The  percent  relative  standard  deviations   (based  on  triplicate
analysis) of  the  relative   response  factors  for  each calibration standard
solution should not  exceed 15  percent.     If this condition is not satisfied,
remedial action should  be taken.

     6.8  The Laboratory must  not  proceed  with  analysis  of samples before
determining and documenting acceptable calibration with the criteria  specified
in Paragraphs 6.3 and 6.7.

     6.9  Routine calibration;

          6.9.1   Inject a  2-uL  aliquot   of  the   column  performance  check
     mixture.  Acquire  at least five data   points for  each GC peak and use the
     same data acquisition time for each of the ions being monitored.
          NOTE:   The same  data  acquisition  parameters  previously  used to
                  analyze concentration  calibration  solutions  during initial
                  calibration  must be used   for the performance check  solution.
                  The column performance  check  solution  must  be  run at the
                  beginning and end  of  a   12  hr  period.   If the contractor
                  laboratory    operates   during   consecutive   12-hr periods
                  (shifts), analysis of the  performance  check solution at the
                  beginning of each 12-hr period and  at the end of  the final
                  12-hr  period is  sufficient.

     Determine and   document   acceptable  column  performance  as  described in
     Paragraph 6.3.

          6.9.2   Inject a 2-uL aliquot of the calibration  standard solution at
     500  ng/mL at the   beginning  of   a  2-hr  period.  Determine  and document
     acceptable   calibration   as  specified  in    Paragraph   6.3,   i.e.,  SIM
     sensitivity  and relative ion abundance  criteria.  The measured RRF's of
                                   8280 - 9
                                                          Revision       0
                                                          Date  September 1986

-------
     all  analytes must be within +30 percent of the mean  values  established by
     Initial  analyses of the calibration standard solutions.


7.0  QUALITY CONTROL

     7.1  Before processing any samples,  the analyst must demonstrate through
the  analysis  of  a  method  blank   that  all  glassware  and   reagents  are
interferent-free at the  method  detection  limit  of  the matrix of interest.
Each time a set of samples is  extracted,  or there is a  change  in reagents,  a
method  blank  must   be   processed   as   a   safeguard  against  laboratory
contamination.

     7.2  A laboratory "method blank" must  be  run along with each analytical
batch (20 or fewer samples).  A  method blank is performed by executing all  of
the specified extraction and cleanup  steps,  except for the introduction of a
sample.  The method blank  is  also  dosed  with  the Internal standards.  For
water samples, one liter of deionized and/or distilled water should be used as
the method blank.  Mineral  oil  may  be  used  as  the method blank for other
matrices.

     7.3  The laboratory will  be   expected  to analyze performance evaluation
samples as provided by the EPA on   a periodic basis throughout the course of a
given project.   Additional  sample analyses  will  not  be  permitted if the
performance criteria  are not achieved.    Corrective  action must be taken and
acceptable performance must be demonstrated before  sample analyses can resume.

     7.4  Samples may be split  with  other  participating  labs on a periodic
basis to ensure  interlaboratory consistency.   At   least one sample per set of
24 must be run  1n duplicate to determine  intralaboratory precision.

     7.5  Field  duplicates  (individual  samples taken  from the same location at
the  same  time)  should  be   analyzed   periodically  to  determine  the total
precision  (field and  lab).

     7.6  Where  appropriate, "field blanks"  will   be  provided to monitor for
possible cross-contamination of   samples  in   the   field.   The typical  "field
blank" will  consist  of  uncontaminated soil  (background soil taken off-site).

     7.7  GC  column  performance   must   be  demonstrated  initially and verified
prior to analyzing  any  sample  in   a  12-hr  period.   The GC column performance
check solution   must  be   analyzed  under  the  same  chromatographic and mass
spectrometric conditions  used  for other samples  and standards.

     7.8  Before using   any cleanup procedure,   the  analyst  must process a
series of  calibration   standards  (Paragraph  6.2)   through  the procedure to
validate elution patterns  and  the absence of  interferents  from  reagents.  Both
alumina  column  and  carbon  column  performance  must  be  checked.   Routinely check
the  8 percent CH2Cl2/hexane eluate  of  environmental  extracts from the alumina
column  for  presence of  target  analytes.
     NOTE:   This fraction  is intended to contain  a high  level  of  interferents
             and analysis  near  the method detection limit may  not be possible.


                                   8280  - 10
                                                          Revision      0
                                                          Date  September 1986

-------
8.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

     8.1  Grab and composite samples  must  be  collected in glass containers.
Conventional sampling practices must  be  followed.    The  bottle must not be
prewashed  with  sample  before  collection.    Composite  samples  should  be
collected in glass containers.    Sampling  equipment  must  be free of tygon,
rubber tubing, other potential sources  of  contamination which may absorb the
target analytes.

     8.2  All samples must be  stored  at  4*C,  extracted  within 30 days and
completely analyzed within 45 days of collection.


9.0  EXTRACTION AND CLEANUP PROCEDURES

     9.1  Internal standard addition.  Use a sample aliquot of 1 g to 1,000 mL
(typical sample size requirements  for  each  type  of  matrix are provided in
Paragraph 9.2) of the chemical  waste  or  soil  to be analyzed.  Transfer the
sample to a tared  flask  and  determine  the  weight  of  the sample.  Add an
appropriate quantity of 13Ci2-2,3,7,8-TCDD, and any other material which is to
be used as an  internal  standard,   (Paragraph  6.2).    All samples should be
spiked with at least  one  internal  standard, for example, 13Ci2-2,3,7,8-TCDD,
to give a concentration of 500 ng/mL in the final concentrated extract.  As an
example, a  10 g sample concentrated  to  a  final volume of 100 uL requires the
addition of 50 ng of 13Ci2-2,3,7,8-TCDD,  assuming 100% recovery.  Adoption of
different  calibration  solution   sets   (as   needed  to  achieve  different
quantification limits  for different   congeners)  will  require a change In the
fortification level.    Individual   concentration  levels  for each  homologous
series must be specified.

     9.2  Extraction

          9.2.1   Sludge/fuel  oil.  Extract  aqueous sludge  samples by refluxing
     a  sample  (e.g. 2  g) with 50  mL  of   toluene  (benzene) in a 125-mL  flask
     fitted with  a Dean-Stark water   separator.  Continue  refluxing  the sample
     until  all the  water  has  been  removed.     Cool  the sample,  filter the
     toluene  extract through   a   fiber  filter,  or   equivalent,  into  a 100-mL
     round  bottom flask.  Rinse the  filter  with 10 mL of  toluene, combine the
     extract  and  rinsate.  Concentrate  the   combined solution to near dryness
     using  a  rotary evaporator at  50*C.     Use   of an inert gas to concentrate
     the extract  is also permitted.   Proceed  with  Step 9.2.4.

          9.2.2   Still bottom.     Extract   still   bottom   samples  by  mixing  a
     sample  (e.g., 1.0 g) with 10  mL  of   toluene  (benzene) in a small beaker
     and filtering the solution through  a  glass  fiber filter  (or equivalent)
     Into a  50-mL round bottom flask.    Rinse the  beaker and filter  with  10 mL
     of toluene.  Concentrate the  combined   toluene solution to near dryness
     using  a  rotary evaporator at  50*C  while  connected to a water  aspirator.
     Proceed with Step 9.2.4.
                                   8280 -  11
                                                          Revision      0
                                                          Date   September  1986

-------
     9.2.3  Fly ash.  Extract fly  ash  samples by placing a sample (e.g.
10 g) and an equivalent amount  of  anhydrous sodium sulfate in a Soxhlet
extraction apparatus charged with 100 ml of toluene (benzene) and extract
for 16 hr using a three cycle/hour schedule.  Cool and filter the toluene
extract through a glass  fiber  filter  paper  into a 500-mL round bottom
flask.  Rinse the filter with 5  ml of toluene.  Concentrate the combined
toluene solution to  near  dryness  using  a  rotary  evaporator at 50*C.
Proceed with Step 9.2.4.

     9.2.4  Transfer the  residue  to  a  125-mL  separatory funnel using
15 ml of hexane.  Rinse  the  flask  with two 5-mL aliquots of hexane and
add the rinses to  the  funnel.    Shake  2  min  with  50  ml of 5% NaCl
solution, discard the aqueous layer and proceed with Step 9.3.

     9.2.5  Soil.  Extract soil samples by placing the sample (e.g. 10 g)
and  an  equivalent  amount  of  anhydrous  sodium  sulfate  in  a 500-mL
Erlenmeyer flask fitted with a Teflon stopper.  Add 20 ml of methanol and
80 ml of petroleum ether, in that order, to the flask.  Shake on a wrist-
action shaker for two hr.  The solid portion of sample should mix freely.
If a smaller soil  aliquot  is  used,  scale  down the amount of methanol
proportionally.

          9.2.5.1  Filter the  extract  from  Paragraph  9.2.5  through a
     glass funnel  fitted  with  a  glass   fiber  filter  and filled with
     anhydrous   sodium   sulfate   into   a   500-mL   Kuderna-Danish   (KD)
     concentrator fitted with  a 10-mL  concentrator   tube.   Add 50 ml. of
     petroleum  ether to  the   Erlenmeyer  flask,  restopper  the flask and
      swirl the  sample gently,  remove the stopper  carefully and decant the
      solvent  through the funnel as  above.   Repeat this procedure with two
      additional  50-mL   aliquots  of petroleum  ether.    Wash the  sodium
      sulfate  in the funnel with two additional  5-mL  portions of petroleum
      ether.

           9.2.5.2   Add  a Teflon  or  PFTE   boiling   chip  and a  three-ball
      Snyder  column  to  the  KD flask.   Concentrate in a 70*C  water bath to
      an  apparent volume of 10 mL.     Remove the  apparatus  from the water
      bath and allow it  to  cool for  5 min.

           9.2.5.3   Add  50  mL of  hexane and   a  new boiling  chip to the KD
      flask.   Concentrate in  a water bath  to an apparent  volume of 10 mL.
      Remove  the apparatus  from the   water   bath  and  allow to cool for  5
      min.

           9.2.5.4  Remove  and invert the Snyder  column  and rinse it down
      into the KD with  two  1-mL  portions  of hexane.  Decant the contents
      of the  KD and  concentrator  tube  into  a 125-mL separatory funnel.
      Rinse the KD with two  additional  5-mL portions of hexane,  combine.
      Proceed with Step 9.3.

      9.2.6  Aqueous samples:  Mark the water  meniscus on the side of the
 1-L sample bottle for  later  determination  of  the exact sample volume.
                              8280 - 12
                                                     Revision      0
                                                     Date  September 1986

-------
    Pour the entire sample  (approximately  1-1} Into a 2-1 separatory funnel.
    Proceed with Step 9.2.6.1.
         NOTE:  A continuous  liquid-liquid extractor may  be used 1n place of
                a separatory  funnel  when  experience  with  a  sample from a
                given source  Indicates that  a  serious emulsion problem will
                result or   an  emulsion  is  encountered  using  a separatory
                funnel.   Add  60  ml  of  methylene  chloride  to the sample
                bottle,  seal,   and   shake  for  30  sec  to  rinse  the Inner
                surface.  Transfer the solvent  to the extractor.  Repeat the
                sample bottle rinse  with an  additional 50- to 100-mL portion
                of methylene  chloride  and   add  the  rinse to the extractor.
                Add 200  to  500  ml   of  methylene  chloride to the distilling
                flask;   add  sufficient   reagent   water  to  ensure  proper
                operation,  and extract for 24 hr.  Allow to cool, then detach
                the distilling flask.    Dry and  concentrate the extract as
                described in  Paragraphs  9.2.6.1  and  9.2.6.2.  Proceed with
                Paragraph 9.2.6.3.

               9.2.6.1  Add  60 mL methylene  chloride  to  the sample bottle,
         seal  and shake  30   sec  to   rinse   the  inner surface.  Transfer the
         solvent to the  separatory funnel  and  extract the sample by shaking
         the  funnel for  2 min with periodic  venting.  Allow the organic layer
         to separate  from the water  phase  for  a  minimum of 10 min.  If the
         emulsion interface between  layers is  more than one-third the volume
         of the solvent  layer, the   analyst  must employ mechanical techniques
         to complete  the phase separation.   Collect the methylene chloride
          (3 x  60  ml)   directly  Into  a  500-mL Kuderna-Danish concentrator
          (mounted with   a   10-mL  concentrator  tube)  by  passing the sample
         extracts through a filter funnel packed with a glass wool plug and
         5 g  of anhydrous sodium sulfate.    After the third extraction, rinse
         the  sodium sulfate with an  additional 30 ml of methylene chloride to
         ensure quantitative transfer.

               9.2.6.2 Attach a Snyder  column  and  concentrate the extract on
          a water  bath until the apparent  volume  of the  liquid  reaches 5 ml.
          Remove the  K-D  apparatus and allow  it to drain and cool  for at least
          10 min.   Remove the Snyder   column,  add 50 ml hexane,  re-attach  the
         Snyder column  and  concentrate   to   approximately  5  ml.    Add a  new
          boiling  chip to the K-D  apparatus  before  proceeding with the  second
          concentration  step.

          Rinse the  flask and the lower  joint with 2 x  5 ml hexane  and combine
          rinses with  extract to give a  final volume of  about  15  mL.

               9.2.6.3  Determine the original  sample  volume  by  refilling  the
          sample bottle  to the mark and   transferring the  liquid  to  a 1,000-mL
          graduated  cylinder.  Record the  sample   volume  to the  nearest  5  ml.
          Proceed  with Paragraph 9.3.

     9.3  In  a 250-mL Separatory funnel,   partition  the  solvent  (15  ml hexane)
against 40  ml  of  20  percent  (w/v)   potassium  hydroxide.    Shake  for 2 m1n.
                                  8280 - 13
                                                         Revision      0
                                                         Date  September 1986

-------
Remove and discard the aqueous layer  (bottom).  Repeat the base washing until
no color is visible in the  bottom  layer  (perform base washings a maximum of
four times).   Strong  base  (KOH)  is  known  to degrade certain PCDD/PCDF's,
contact time must be minimized.

     9.4  Partition the solvent (15  ml  hexane)  against  40  ml of 5 percent
(w/v) sodium chloride.  Shake  for  2  min.   Remove and discard aqueous layer
(bottom).
     NOTE:  Care  should  be  taken  due  to  the  heat  of neutralization and
     hydration.

     9.5  Partition the solvent (15 ml  hexane)  against 40 ml of concentrated
sulfuric acid.   Shake  for  2  min.    Remove  and  discard the aqueous layer
(bottom).  Repeat the acid  washings  until  no  color  is visible in the acid
layer.  (Perform acid washings a maximum of four times.)

     9.6  Partition the  extract  against  40  ml  of  5  percent  (w/v) sodium
chloride.  Shake for 2 min.    Remove  and discard the aqueous layer  (bottom).
Dry  the organic layer by pouring  through a funnel containing anhydrous sodium
sulfate into a 50-mL round bottom  flask,  wash the separatory funnel with two
15-mL portions of hexane,  pour  through  the  funnel,  and combine the hexane
extracts.  Concentrate  the  hexane   solution  to  near  dryness with a rotary
evaporator  (35*C water bath), making  sure  all traces of toluene  are removed.
(Use of  blowdown  with  an  inert  gas  to  concentrate  the  extract is also
permitted).

     9.7  Pack a gravity column  (glass 300-mm  x 10.5-mm), fitted with a Teflon
stopcock, in the following manner:

     Insert a glass-wool plug  into the bottom  of the  column.  Add a  4-g layer
of sodium sulfate.  Add a 4-g  layer of Woelm super 1 neutral alumina.  Tap the
top  of the  column gently.  Woelm  super 1 neutral alumina need not  be  activated
or cleaned  prior to use but  should be stored in a sealed desiccator.  Add a 4-
g layer of  sodium sulfate to cover  the  alumina.   Elute with  10  ml  of hexane
and  close the  stopcock just  prior to  the  exposure of  the  sodium sulfate layer
to air.  Discard  the  eluant.   Check  the column for channeling.  If channeling
is present  discard  the column.   Do not tap  a wetted column.

      9.8   Dissolve  the residue from Step 9.6   in  2 ml of  hexane and  apply  the
hexane  solution  to  the top  of  the column.   Elute with enough  hexane (3-4  ml)
to  complete the  transfer  of the sample   cleanly to the surface  of  the alumina.
Discard  the eluant.

           9.8.1   Elute with 10 ml  of  8  percent  (v/v)  methylene chloride in
      hexane.   Check by  GC/MS analysis  that no PCDD's or PCDF's are  eluted in
      this  fraction.  See Paragraph 9.9.1.

           9.8.2  Elute the PCDD's and PCDF's from  the column with 15 ml of 60
      percent (v/v)  methylene chloride in hexane and  collect this fraction in a
      conical  shaped (15-mL) concentrator tube.
                                   8280 - 14
                                                          Revision      0
                                                          Date  September 1986

-------
    9.9  Carbon column cleanup;

         Prepare a carbon column as described in Paragraph 4.18.

         9.9.1  Using a  carefully  regulated  stream  of nitrogen (Paragraph
    4.15),  concentrate  the  8  percent  fraction  from  the  alumina column
    (Paragraph 9.8.1) to about 1 ml.  Wash the sides of the tube with a small
    volume of hexane  (1 to 2 ml) and  reconcentrate to about 1 ml.  Save this
    8 percent concentrate for  GC/MS  analysis  to  check for breakthrough of
    PCDD's and PCDF's.  Concentrate the 60 percent fraction  (Paragraph 9.8.2)
    to  about 2 to  3   ml.    Rinse  the carbon with 5 ml cyclohexane/methylene
    chloride  (50:50 v/v) 1n the  forward  direction  of  flow and then In the
    reverse direction of flow.  While still  in the reverse direction of flow,
    transfer the sample concentrate to  the  column  and  elute with 10 ml of
    cyclohexane/methylene  chloride   (50:50  v/v)   and  5  ml  of  methylene
    chloride/methanol/benzene  (75:20:5, v/v).    Save  all  above eluates and
    combine  (this  fraction may be  used as a  check on column efficiency).  Now
    turn  the  column over  and  in  the  direction  of  forward  flow elute the
    PCDD/PCDF  fraction with 20 ml  toluene.
          NOTE:   Be sure no carbon  fines are  present in the eluant.

          9.9.2   Alternate carbon column cleanup.  Proceed as  in Section 9.9.1
    to  obtain  the  60  percent  fraction   re-concentrated  to  400 uL which is
    transferred  to an HPLC   injector  loop   (1  ml).    The  injector  loop 1s
    connected to the  optional  column  described   in Paragraph  4.18.  Rinse the
    centrifuge tube with  500  uL  of  hexane  and  add  this   rinsate to the
    injector  loop.   Load  the  combined  concentrate   and   rinsate onto the
    column.   Elute the column  at 2 mL/min, ambient temperature, with 30 ml of
    cyclohexane/methylene chloride 1:1  (v/v).  Discard the eluant.  Backflush
    the column with 40  mL  toluene   to   elute   and collect  PCDD's  and PCDF's
     (entire   fraction).     The  column   is    then  discarded  and   30  ml  of
    cyclohexane/methylene chloride 1:1  (v/v)  is  pumped through a  new column
    to  prepare  it  for the next sample.

          9.9.3   Evaporate the  toluene  fraction  to  about   1   ml on a rotary
    evaporator using  a water  bath  at   50*C.   Transfer to a 2.0-mL React1-vial
    using a  toluene  rinse  and  concentrate  to  the  desired  volume  using  a
    stream of N£.   The  final  volume  should be  100 uL for soil samples  and
    500 uL for sludge,  still   bottom,   and   fly  ash samples;  this is provided
    for guidance,  the correct  volume   will   depend on the  relative  concentra-
    tion of target analytes.   Extracts  which are determined  to  be outside the
    calibration  range for individual   analytes   must  be diluted or a  smaller
    portion  of the sample must be   re-extracted.  Gently swirl  the  solvent on
    the lower portion of   the  vessel   to  ensure complete dissolution of the
     PCDD's and PCDF's.

     9.10  Approximately 1  hr before   HRGC/LRMS  analysis,  transfer  an  aliquot
of the extract to  a  micro-vial   (Paragraph   4.16).     Add  to this  sufficient
recovery standard  (13Ci2l,2,3,4-TCDD)  to  give  a  concentration of  500 ng/mL.
(Example:  36 uL aliquot of  extract   and  4  uL of recovery standard solution.
Remember to adjust the final  result  to  correct for  this dilution.   Inject  an
appropriate aliquot (1 or 2  uL) of the sample into  the  GC/MS  instrument.


                                  8280 - 15
                                                         Revision       0
                                                         Date  September  1986

-------
10.0  GC/MS ANALYSIS

     10.1  When toluene Is employed as the final  solvent use of a bonded phase
column from Paragraph 4.3.2 is  recommended.   Solvent exchange Into trldecane
is required for other liquid phases or nonbonded columns (CP-Sil-88).
     NOTE:  Chromatographic conditions must be adjusted to account for solvent
            boiling points.

     10.2  Calculate response factors for  standards  relative to the internal
standards, 13C12-2,3,7,8-TCDD  and  13Ci2-OCDD  (see  Section  11).    Add the
recovery standard (13Ci2-l,2,3,4-TCDD) to the samples prior to injection.  The
concentration of the recovery standard in  the sample extract must be the same
as that in the calibration standards used to measure the response factors.

     10.3  Analyze samples with selected ion monitoring, using all of the ions
listed in Table 2.  It is recommended  that the GC/MS run be divided into five
selected 1on monitoring sections, namely:   (1) 243, 257,, 304, 306, 320, 322,
332, 334, 340, 356,, 376   (TCDD's,  TCDF's,  13Ci2-labeled internal and recovery
standards, PeCDD's,, PeCDF's, HxCDE);  (2)  277,  293,  306, 332, 338, 340, 342,
354, 356, 358, 410  (peCDD's,  PeCDF's,  HpCDE);   (3)  311, 327, 340, 356, 372,
374, 376, 388, 390,, 392,  446,   (HxCDD's,   HxCDF's,  OCDE);  (4) 345, 361, 374,
390, 406, 408, 410r 422,  424,  426,  480  (HpCDD's, HpCDF's,  NCDE) and (5) 379,
395, 408, 424, 442,,  444,  458,   460,  470,  472, 514  (OCDD, OCDF,  13Ci2-OCDD,
DCDE).   Cycle  time  not  to   exceed   1  sec/descriptor.   It is  recommended that
selected  ion monitoring section   1   should  be  applied  during   the GC  run to
encompass the  retention window  (determined  in  Paragraph  6.3) of the  first- and
last-eluting tetra-chlorinated isomers.   If a  response  is observed  at m/z 340
or 356,  then the  GC/MS   analysis must   be repeated;  selected ion  monitoring
section  2  should  then  be applied   to  encompass the  retention  window  of the
first- and  last-eluting penta-chlorinated isomers.    HxCDE,  HpCDE,  OCDE,  NCDE,
DCDE, are  abbreviations for   hexa-,   hepta-, octa-,  nona-, and decachlorinated
diphenyl  ether,  respectively.

      10.4   Identification criteria for PCDD's  and PCDF's:

           10.4.1  All  of  the   characteristic  ions,   i.e.  quantisation ion,
      confirmation ions,  listed  in Table   2  for   each  class of  PCDD and PCDF,
      must  be present  in the   reconstructed  ion  chromatogram.   It is desirable
      that  the  M  -  COC1  ion   be   monitored   as an  additional requirement.
      Detection limits  will  be based   on quantitation ions within  the molecules
      in  cluster.

           10.4.2  The  maximum  intensity   of   each   of  the   specified  charac-
      teristic  ions  must  coincide within 2 scans  or 2 sec.

           10.4.3  The  relative  intensity  of the  selected,  isotopic  ions  within
      the molecular  ion cluster  of a  homologous series of PCDD's  of PCDF's must
      lie within  the range specified  in  Table  3.

           10.4.4  The GC  peaks  assigned  to a  given  homologous  series must have
      retention times within  the  window   established  for  that series by the
      column performance  solution.


                                   8280 - 16
                                                          Revision      0
                                                          Date  September 1986

-------
     10.5  Quantitate the PCDD and  PCDF  peaks  from the response relative to
the appropriate Internal  standard.    Recovery  of each Internal  standard)  vs.
the recovery standard must be greater than 40 percent.  It Is recommended that
samples with recoveries of less than 40 percent or greater than 120 percent be
re-extracted and re-analyzed.
     NOTE:  These  criteria  are  used  to  assess  method  performance;   when
            properly applied, Isotope  dilution  techniques are Independent of
            Internal standard recovery.

In those  circumstances  where  these  procedures  do  not  yield a definitive
conclusion, the use  of  high  resolution  mass  spectrometry or HRGC/MS/MS is
suggested.


11.0  CALCULATIONS

      NOTE:  The relative response  factors  of  a  given  congener within any
             homologous series  are  known  to  be  different.    However, for
             purposes of these  calculations,  1t  will  be assumed that every
             congener within a  given  series  has  the same relative response
             factor.  In order to  minimize  the  effect of this assumption on
             risk   assessment,   a   2,3,7,8-substituted   isomer   that   1s
             commercially  available  was  chosen  as  representative  of each
             series.  All relative  response   factor  calculations for a given
             homologous series are based on that compound.

      11.1   Determine the concentration of  individual  isomers of tetra-, penta,
and hexa-CDD/CDF according to the equation:


           Concentration, ng/g  =  G        X$RRF
 where :
      Qis   =   "9  of  internal  standard   13Ci2-2,3,7,8-TCDD,  added to the sample
              before extraction.

        G   =   g of sample  extracted.

       As   =   area of quantisation  ion  of  the compound of  interest.

      AJS   =   area of quantitatlon   1on  (m/z   334)  of  the  internal  standard,
              13C12-2,3,7,8-TCDD.

      RRF   =   response factor  of   the  quantitation  ion  of the  compound of
              Interest relative to  m/z  334 of 13Ci2-2,3,7,8-TCDD.

      NOTE:   Any  dilution   factor   introduced   by  following  the  procedure in
             Paragraph 9.10 should  be applied to this calculation.
                                   8280 - 17
                                                         Revision      0
                                                         Date  September 1986

-------
          11.1.1  Determine the concentration of  individual  isomers  of  hepta-
     CDD/CDF and the concentration of OCDD and OCDF according to  the  equation:

                                    Qis x As
          Concentration, ng/g  =  G x     x RRF
where :
     Qis  =  n9 °f internal standard    c^-OCDD,   added  to the  sample before
             extraction.

       G  =  g of sample extracted.

      AS  =  area of quantitation ion of the compound of interest.

     Ajs  =  area of quantitation  ion  (m/z  472)   of  the internal  standard,
             13Ci2-OCDD.

     RRF  =  response factor  of  the  quantitation  ion  of  the  compound of
             interest relative to m/z 472 of 13Ci2-OCDD.

     NOTE:  Any dilution   factor  introduced  by  following  the  procedure in
            Paragraph 9.10 should be applied to this calculation.

          11.1.2  Relative response factors are calculated using data obtained
     from the analysis of  multi -level  calibration standards according to the
     equation:
           RRF  =  ^" ^   C
                   is     s

 where:

      As   =  area  of quantitation  ion  of the compound of interest.

     AIS   =  area  of quantitation   ion of  the  appropriate internal standard
              (m/z  334  for  13C12-2,3,7,8-TCDD; m/z 472 for 13C12-OCDD).

      C^s   =  concentration  of the  appropriate internal  standard,
              l3Ci2-2,3,7,8-TCDD or 13C12-OCDD)

      Cs   =  concentration  of the  compound  of interest.

           11.1.3  The  concentrations   of  unknown   isomers  of  TCDD  shall be
      calculated  using  the  mean RRF determined for 2,3,7,8-TCDD.

           The  concentrations of unknown   isomers  of PeCDD  shall  be calculated
      using the  mean   RRF   determined   for   1,2,3,7,8-PeCDD or   any  available
      2,3,7,8,X-PeCDD  isomer.
                                   8280 - 18
                                                          Revision
                                                          Date   September  1986

-------
     The concentrations of unknown  isomers  of HxCDD shall  be calculated
using the mean  RRF  determined  for  1,2,3,4,7,8-HxCDD  or any available
2,3,7,8,-X,Y-HXCDD isomer.

     The concentrations of unknown  isomers  of HpCDD shall  be calculated
using the mean RRF  determined  for  1,2,3,4,6,7,8-HpCDD or any available
2,3,7,8,X,YrZ-HpCDD isomer.

     The concentrations of unknown  isomers  of  TCDF shall  be calculated
using the mean RRF determined for 2,3,7,8-TCDF.

     The concentrations of unknown  isomers  of PeCDF shall  be calculated
using the  mean  RRF  determined  for  1,2,3,7,8-PeCDF  or  any available
2,3,7,8lX-PeCDF isomer.

     The concentrations of unknown  isomers  of HxCDF shall  be calculated
using the  mean  RRF  determined  for  1,2,4,7,8-HxCDF  or  any available
2,3,7,8-X.Y-HxCDF isomer.

     The concentrations of unknown  isomers  of HpCDF shall  be calculated
using the mean RRF  determined  for  1,2,3,4,6,7,8-HpCDF or any available
2I3,7,8lX,Y,Z-HpCDF isomer.

     The concentration of the  octa-CDD  and octa-CDF shall  be calculated
using the mean RRF determined for each.

     Mean relative response factors  for  selected  PCDD's and PCDF's are
given in Table 4.

     11.1.4   Calculate  the  percent  recovery,  Rjs,  for  each internal
standard in the sample extract, using the equation:

              Ais  x Q
     R
       1s    Ax RFr xQ.s

where:

     Ars   =  Area of quantisation ion  (m/z 334) of the recovery standard,
             13Ci2-l,2,3,4-TCDD.

     Qrs   =  "9  °f  recovery   standard,  13Ci2-l,2,3,4-TCDD,  added  to
             extract.

The  response factor  for   determination  of  recovery is calculated using
data obtained  from the  analysis  of the multi-level calibration standards
according  to the equation:

     RF  _ Ais  x  Crs
     RFr~  ArsxCis
                              8280 -  19
                                                    Revision
                                                    Date  September 1986

-------
     where:

          Crs  =  Concentration of the recovery standard,  13Ci2-l,2,3,4-TCDD.

          11.1.5  Calculation of  total   concentration   of  all  Isomers  within
     each homologous series of PCDD's and PCDF's.

     Total concentration  =  Sum of the concentrations  of the Individual
     of PCDD's or PCDF's     PCDD or PCDF Isomers

     11.4  Report results In  nanograms  per  gram;   when duplicate and  spiked
samples are reanalyzed, all data obtained should be  reported.

     11.5  Accuracy and Precision.    Table  5  gives  the  precision data for
revised Method 8280 for  selected  analytes  in the matrices shown. Table 6
lists recovery data for the same analyses.  Table  2  shows the linear range and
variation of  response  factors  for  selected  analyte  standards.   Table 8
provides the method detection limits as measured in  specific sample matrices.

     11.6  Method Detection  Limit.    The  Method  Detection  Limit  (MDL) is
defined as the minimum concentration of  a  substance that can be measured and
reported with 99  percent  confidence  that  the  value  1s  above  zero.  The
procedure used to determine the  MDL  values  reported  in Table 8 was obtained
from Appendix  A  of  EPA  Test  Methods  manual,   EPA-600/4-82-057 July 1982,
"Methods  for  Organic   Chemical   Analysis   of   Municipal  and  Industrial
Wastewater."

     11.7  Maximum Holding Time  (MHT).   Is  that   time  at which a 10 percent
change in the analyte  concentration  (Ctio)  occurs  and the precision  of the
method of  measurement  allows   the  10  percent  change  to  be statistically
different from the 0 percent change   (Cto) at the 90 percent confidence level.
When  the  precision  of   the    method   is  not  sufficient  to  statistically
discriminate a 10 percent  change in  the concentration from 0 percent  change,
then the maximum holding time   is  that  time  where the percent change in the
analyte concentration  (Ctn)  is   statistically different than the concentration
at 0 percent change  (Cto)  an(*  greater than 10 percent change at the 90  percent
confidence level.
                                   8280 - 20
                                                          Revision       0
                                                          Date   September 1986

-------
TABLE 1.  REPRESENTATIVE GAS CHROMATOGRAPH RETENTION TIMES* OF ANALYTES
Analyte
2,3,7,8-TCDF
2,3,7,8-TCDD
1,2,3, 4-TCDD
1,2,3,4,7-PeCDD
1,2,3,4,7,8-HxCDD
1,2,3,4,6,7,8-HpCDD
OCDD
50-m
CP-S11-88
25.2
23.6
24.1
30.0
39.5
57.0
NM
30-m
DB-5
17.8
17.4
17.3
20.1
22.1
24.1
25.6
3— m
SP-2250
26.7
26.7
26.5
28.1
30.6
33.7
NM
*Retention time in min, using temperature programs shown below.
NM  =  not measured.
Temperature Programs;
          CP-Sil-88           60*C-190*C at 20'/nrin; 190*-240* at 5*/min.
          DB-5                170*, 10 min; then at 8*/min to 320'C, hold
          30 m x 0.25 mm      at 320*C 20 min  (until OCDD elutes).
          Thin film  (0.25 urn)
          SP-2250             70'-320' at lO'/minute.
                            Column Manufacturers
CP-Sil-88            Chrompack,  Incorporated, Bridgewater, New Jersey
DB-5,                J  and   W   Scientific,   Incorporated,  Rancho  Cordova,
                     California
SP-2250              Supelco,     Incorporated,     Bellefonte,    Pennsylvania
                                   8280 -  21
                                                          Revision
                                                          Date   September  1986

-------
                    TABLE 2.   IONS SPECIFIED3 FOR SELECTED ION MONITORING
                            FOR PCDD'S AND PCDF'S
                        Quantitation
                            ion
                    Confirmation
                        ions
                      M-COC1
PCDD's

13Ci2-Tetra
Tetra
Penta
Hexa
Hepta
Octa
13Ci2-Octa

PCDF's
334
322
356
390
424
460
472
332
320
354;358
388;392
422;426
458
470
257
293
327
361
395
Tetra
Penta
Hexa
Hepta
Octa
306
340
374
408
444
304
338; 342
372; 376
406;410
442
243
277
311
345
379
alons at m/z  376  (HxCDE), 410  (HpCDE), 446  (OCDE), 480  (NCDE)  and  514  (DCDE)
 are also  included  in the scan monitoring sections  (1)  to  (5),  respectively.
 See Paragraph  10.3.
  TABLE  3.   CRITERIA  FOR ISOTOPIC  RATIO  MEASUREMENTS  FOR  PCDD'S  AND  PCDF'S

                         Selected ions  (m/z)                  Relative intensity
 PCDD's

 Tetra
 Penta
 Hexa
 Hepta
 Octa

 PCDF's
  320/322
  358/356
  392/390
  426/424
  458/460
              0.65-0.89
              0.55-0.75
              0.69-0.93
              0.83-1.12
              0.75-1.01
Tetra
Penta
Hexa
Hepta
Octa
304/306
342/340
376/374
410/408
442/444
0.65-0.89
0.55-0.75
0.69-0.93
0.83-1.12
0.75-1.01
                                   8280 - 22
                                                          Revision      0
                                                          Date  September 1986

-------
     TABLE 4.  MEAN  RELATIVE  RESPONSE FACTORS OF CALIBRATION STANDARDS
Analyte
2,3,7,8-TCDD
1,2,3,7,8-PeCDD
1,2,3,4,7,8-HxCDD
l,2,3,4,6,7,8-HpCDDb
OCDDb
2,3,7,8-TCDF
1,2,3,7,8-PeCDF
1,2,3,4,7,8-HxCDF
l,2,3,4,6,7,8-HpCDFb
OCDFb
13Ci2-2,3,7,8-TCDD
13Ci2-l,2,3,4-TCDD
13Ci2-OCDD
RRFa
1.13
0.70
0.51
1.08
1.30
1.70
1.25
0.84
1.19
1.57
1.00
0.75
1.00
RSD%
(n = 5)
3.9
10.1
6.6
6.6
7.2
8.0
8.7
9.4
3.8
8.6
-
4.6
-
Quantitation ton
(m/z)
322
356
390
424
460
306
340
374
444
408
334
334
472
aThe RRF value is the mean of the five determinations made.  Nominal weights
 injected were 0.2, 0.5, 1.0, 2.0 and 5.0 ng.

bRRF values for these analytes were determined relative to 13Cj2-OCDD.  All
 other RRF's were determined relative to 13Ci2-2,3,7,8-TCDD.

Instrument Conditions/Tune - GC/MS system was tuned as specified in
                             Paragraph 6.3.  RRF data was acquired under
                             SIM control, as specified in Paragraph 10.3.

GC Program - The GC column temperature was programmed as specified in
             Paragraph 4.3.2(b).
                                  8280 - 23
                                                         Revision
                                                         Date  September 1986

-------
TABLE 5.  PRECISION DATA FOR REVISED METHOD 8280
Compound
2,3,7,8-TCDD




1,2,3,4-TCDD




1,3,6,8-TCDD




1,3,7,9-TCDD




1,3,7,8-TCDD




1,2,7,8-TCDD




1,2,8,9-TCDD




Analyte
Matrix3
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
level (ng/g)
Native
NDb
378
ND
ND
487
ND
ND
ND
38.5
ND
ND
ND
ND
19.1
227
ND
ND
ND
58.4
ND
ND
ND
ND
16.0
422
ND
ND
ND
2.6
ND
ND
ND
ND
ND
ND
Native
+ spike
5.0
378
125
46
487
5.0
25.0
125
38.5
2500
2.5
25.0
125
19.1
2727
2.5
25.0
125.0
58.4
2500
5.0
25.0
125
16.0
2920
5.0
25.0
125
2.6
2500
5.0
25.0
125
46
2500
N
4
4
4
2
4
3
4
4
4
4
4
4
4
2
2
4
4
4
2
2
4
4
4
4
2
4
4
4
3
2
4
4
4
2
2
Percent
RSD
4.4
2.8
4.8
-
24
1.7
1.1
9.0
7.9
-
7.0
5.1
3.1
-
-
19
2.3
6.5
-
-
7.3
1.3
5.8
3.5
-
7.7
9.0
7.7
23
-
10
0.6
1.9
-
-
                     8280 - 24
                                            Revision      0
                                            Date  September 1986

-------
TABLE 5  (Continued)
Compound
1,2,3,4,7-PeCDD




1,2,3,7,8-PeCDD




1,2,3,4,7,8-HxCDD




1,2,3,4,6,7,8-HpCDD




1,2,7,8-TCDF




1,2,3,7, 8-PeCDF




1,2,3,4,7,8-HxCDF




Analyte
Matrix3
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge0
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom3
clay
soil
sludge
fly ash
still bottom
level (ng/g)
Native
ND
ND
ND
25.8
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
8760
ND
ND
ND
ND
ND
7.4
ND
ND
ND
ND
ND
25600
ND
ND
13.6
24.2
ND
Native
+ spike
5.0
25.0
125
25.8
2500
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
5.0
25.0
8780
-
-
5.0
25.0
125
7.4
2500
5.0
25.0
125
46
28100
5.0
25.0
139
24.2
2500
N
4
4
4
2
2
4
4
4
2
2
4
4
4
2
2
4
4
4
_
-
4
4
4
3
2
4
4
4
2
2
4
4
4
4
2
Percent
RSD
10
2.8
4.6
6.9
-
25
20
4.7
_
-
38
8.8
3.4
-
-
^
_
-
_
-
3.9
1.0
7.2
7.6
-
6.1
5.0
4.8
—
-
26
6.8
5.6
13.5
-
      8280 - 25
                             Revision      0
                             Date  September 1986

-------
                            TABLE 5.   (Continued)


Compound
OCDF




Analyte

Matr1xa
clay
soil
sludge
fly ash
still bottom
level (ng/g)
Native Percent
Native + spike N RSD
ND -
ND -
192 317 4 3.3
ND -
ND -
amatrix types:

 clay:  pottery clay.

 soil:  Times Beach,  Missouri,  soil  blended  to  forma homogeneous sample.
This sample was analyzed as  a  performance evaluation sample for the Contract
Laboratory Program  (CLP)  in  April  1983.    The  results  from EMSL-LV and 8
contract laboratories using  the  CLP  protocol  were  305.8 ng/g 2,3,7,8-TCDD
with a standard deviation of 81.0.

 fly ash:  ash from a municipal incinerator; resource recovery ash No. 1.

 still bottom:  distillation bottoms (tar) from 2,4-dichlorophenol production.

sludge:    sludge   from  cooling   tower  which  received  both  creosote  and
pentachl orophenol 1 c wastewaters .

Cleanup of clay, soil and fly ash samples was through alumina column only.
(Carbon column not  used.)
    - not detected at concentration  Injected  (final volume 0.1 mL or greater).

GEstimated  concentration  out of calibration range of standards.
                                   8280 - 26
                                                          Revision       0
                                                          Date   September 1986

-------
TABLE 6.  RECOVERY DATA FOR REVISED METHOD 8280
Compound
2,3,7,8-TCDD




1,2,3,4-TCDD




1,3,6,8-TCDD




1,3,7,9-TCDD




1,3,7,8-TCDD




1,2,7,8-TCDD




1,2,8,9-TCDD




Matr1xa
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
Nat1veb
(ng/g)
ND
378
ND
ND
487
ND
ND
ND
38.5
ND
ND
ND
ND
19.1
227
ND
ND
ND
58.4
ND
ND
ND
ND
16.0
615
ND
ND
ND
2.6
ND
ND
ND
ND
ND
ND
Sp1kedc
level
(ng/g)
5.0
-
125
46
-
5.0
25.0
125
46
2500
2.5
25.0
125
46
2500
2.5
25.0
125
46
2500
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
Mean
percent
recovery
61.7
-
90.0
90.0
-
67.0
60.3
73.1
105.6
93.8
39.4
64.0
64.5
127.5
80.2
68.5
61.3
78.4
85.0
91.7
68.0
79.3
78.9
80.2
90.5
68.0
75.3
80.4
90.4
88.4
59.7
60.3
72.8
114.3
81.2
                   8280 - 27
                                          Revision      0
                                          Date  September 1986

-------
TABLE 6.  (Continued)
Compound
1,2,3,4,7-PeCDD




1,2,3,7,8-PeCDD




1,2,3,4, 7,8-HxCDD




1,2,3,4,6,7,8-HpCDD




2,3,7,8-TCDD
(C-13)



1,2,7, 8-TCDF




1,2,3,7,8-PeCDF




Matrix3
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soi 1
sludge"
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
Nativeb
(ng/g)
ND
ND
ND
25.8
NO
NO
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
8780
ND
ND
ND
ND
ND
ND
ND
ND
ND
ND
7.4
ND
ND
ND
ND
ND
25600
Spikedc
level
(ng/g)
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
5.0
25.0
125
-
-
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
5.0
25.0
125
46
2500
Mean
percent
recovery
58.4
62.2
79.2
102.4
81.8
61.7
68.4
81.5
104.9
84.0
46.8
65.0
81.9
125.4
89.1
ND
ND

-
-
64.9
78.8
78.6
88.6
69.7
65.4
71.1
80.4
90.4
104.5
57.4
64.4
84.8
105.8
-
       8280 - 28
                              Revision      0
                              Date  September 1986

-------
                            TABLE 6.   (Continued)

Compound
1,2,3,4,7,8-HxCDF




OCDF





Matrixa
clay
soil
si udge
fly ash
still bottom
clay
soil
sludge
fly ash
still bottom
Nativeb
(ng/g)
ND
ND
13.6
24.2
ND
ND
ND
192
ND
ND
Spiked0
level
(ng/g)
5.0
25.0
125
46
2500
_
-
125
-
—
Mean
percent
recovery
54.2
68.5
82.2
91.0
92.9
_
-
86.8
-
-
amatrix types:

clay:  pottery clay.

soil:  Times Beach, Missouri soil blended  to form a homogeneous sample.  This
sample was  analyzed  as  a  performance  evaluation  sample  for the Contract
Laboratory Program (CLP)  in  April  1983.    The  results  from EMSL-LV and 8
contract laboratories using  the  CLP  protocol  were  305.8 ng/g 2,3,7,8-TCDD
with a standard deviation of 81.0.

fly ash:  ash from a municipal incinerator:  resource recovery ash No. 1.

still bottom:  distillation bottoms (tar) from 2,4-dichlorophenol production.

sludge:    sludge  from  cooling   tower  which  received  both  creosote  and
pentachlorophenol wastewaters.

The clay, soil and fly ash  samples  were subjected to alumina column cleanup,
no carbon column was used.
       volume of concentrate 0.1 mL  or greater, ND means below quantification
limit, 2 or more samples analyzed.

cAmount of analyte added to sample, 2 or more samples analyzed.

^Esti mated concentration out of calibration range of standards.
                                  8280 - 29
                                                         Revision      0	
                                                         Date  September 1986

-------
          TABLE 7.   LINEAR RANGE AND  VARIATIOIN OF  RESPONSE FACTORS
Analyte Linear range tested (pg) n&
l,2,7,8-TCDFa
2,3,7,8-TCDDa
2,3,7,8-TCDF
50-6000
50-7000
300-4000
8
7
5
Mean RF
1.634
0.721
2.208
%RSD
12.0
11.9
7.9
Response factors for these analytes were calculated using  2,3,7,8-TCDF  as  the
internal standard.  The response  factors for 2,3,7,8-TCDF  were calculated  vs.
13C12-1,2,3,4-TCDD,

^Each value of n represents a different concentration level.
                                   8280 -  30
                                                         Revision
                                                         Date  September 1986

-------
   TABLE 8.  METHOD DETECTION  LIMITS OF   C12 - LABELED  PCDD'S  and  PCDF'S

             IN  REAGENT  WATER  (PPT) AND  ENVIRONMENTAL  SAMPLES (PPB)
13C -Labeled
Analyte
2,3,7,8-TCDD
1,2,3,7,8-PeCDD
1,2,3,6,7,8-HxCDD
1,2,3,4,6,7,8-HpCDD
OCDD
2,3,7,8-TCDF
1,2,3,7,8-PeCDF
1,2,3,4,7,8-HxCDF
Reagent
Water3
0.44
1.27
2.21
2.77
3.93
0.63
1.64
2.53
Missouri
Soil5
0.17
0.70
1.25
1.87
2.35
0.11
0.33
0.83
Flyg
Ash
0.07
0.25
0.55
1.41
2.11
0.06
0.16
0.30
Industrial
Sludge0
0.82
1.34
2.30
4.65
6.44
0.46
0.92
2.17
Still-
Bottom
1.81
2.46
6.21
4.59
10.1
0.26
1.61
2.27
Fuel
Oil4
0.75
2.09
5.02
8.14
23.2
0.48
0.80
2.09
Fuel C
Sawdus
0.13
0.18
0.36
0.51
1.48
0.40
0.43
2.22
a
.Sample size 1 ,000 mL.
 Sample size 10 g.
.Sample size 2 g.
 Sample size 1 g.
Note:  The final sample-extract volume was 100 uL for all samples.

Matrix types used in MDL Study:

     - Reagent water:  distilled, deionized laboratory water.
     - Missouri soil:  soil blended to form a homogeneous sample.
     - Fly-ash:  alkaline ash recovered from the electrostatic precipitator of
       a coal-burning power plant.
     - Industrial sludge:  sludge from  cooling tower which received creosotic
       and pentachlorophenolic wastewaters.  Sample  was ca. 70 percent water,
       mixed with oil and sludge.
     - Still-bottom:    distillation  bottoms  (tar)  from  2,4-dichlorophenol
       production.
     - Fuel oil:  wood-preservative solution from the modified Thermal Process
       tanks.  Sample  was  an  oily  liquid  (>90  percent oil) containing no
       water.
     - Fuel oil/Sawdust:  sawdust was obtained  as a very fine powder from the
       local lumber yard.   Fuel  oil  (described  above)  was  mixed at the 4
       percent (w/w) level.

Procedure used for the Determination  of  Method Detection Limits was obtained
from "Methods  for  Organic  Chemical  Analysis  of  Municipal  and Industrial
Wastewater" Appendix A, EPA-600/4-82-057,  July  1982.   Using this procedure,
the method detection  limit  is  defined  as  the  minimum  concentration of a
substance that can be measured  and  reported  with 99 percent confidence that
the value is above zero.
                                  8280 - 31
                                                         Revision      0	
                                                         Date  September 1986

-------
 aaoi-8're'z
   dd01-87.'C'3
                                                             0>
                                                             en

                                                             I
                                                             10


                                                             §
                                                             O
                                                             U
                                                             o
                                                             I
                                                             s
                                                             s
                                                             2
                                                             o
                                                             at
CM
3
CD
                                                             U.
 I
O
                       8280 - 32
                                           Revision     o
                                           Date  September 1986

-------
                             METHOD B3BO

POLYCHLOHINATEO OIBENZO-P-DIOXIINS AND POLYCHLORINATEO OI8ENZONFUHANS
f     Start      J
  e. i
 Perform Initial
 calibration on
  GC/MS system
  6.9
                                                      1O.2
     Calculate
     response
   factors for
    standards
   Oo routine
   calibration
                                                      10.3
      Analyze
   samples with
   selected Ion
    monitoring
  9.2
        Extract
    •ample using
    appropriate
  method for the
   waste matrix
  9.9

       Prepare
  carbon column;
     do carbon
  column cleanup
                                                     10.5
Quantltate PCOD
 and PCDF peaks
                                               Yes
    o
 11.01

    Determine
 concentrations
   •nd  report
    results
                                                   f     Stop      J
                                  8280 -  33
                                                             Revision       Q
                                                             Date   September  1986

-------
                                 APPENDIX A

                    SIGNAL-TO-NOISE DETERMINATION METHODS
MANUAL DETERMINATION

     This method corresponds to a manual determination of the S/N from a GC/MS
signal, based on the measurement of  its  peak height relative to the baseline
noise.  The procedure is composed of  four steps as outlined below.  (Refer to
Figure 1 for the following discussion).

     1.   Estimate the peak-to-peak noise (N) by tracing the two lines (Ej and
          £2) defining the noise envelope.   The lines should pass through the
          estimated statistical mean  of  the  positive  and the negative peak
          excursions as shown in Figure 1.  In addition, the signal offset (0)
          should be set high enough such that negative-going noise (except for
          spurious negative spikes) is recorded.

     2.   Draw the  line   (C)  corresponding  to  the  mean  noise between the
          segments defining the noise envelope.

     3.   Measure the height of the GC/MS  signal   (S) at the apex of the peak
          relative to the  mean noise C.   For noisy GC/MS signals, the average
          peak height should be measured from the estimated mean apex signal D
          between £3 and £4.

     4.   Compute the S/N.

     This method of  S/N   measurement   is  a  conventional, accepted method of
noise  measurement in analytical chemistry.


INTERACTIVE  COMPUTER GRAPHICAL METHOD

     This method calls  for the  measurement  of  the  GC/MS peak area using the
computer data system and Eq. 1:
                                    A/t
                         S/N = Aj/2t +  Af/2t
 where  t  is  the  elution  time  window   (time   interval, t2~t2, at the base of the
 peak used to  measure  the  peak  area A).    (Refer to  Figure 2, for the following
 discussion).

     AI  and Ar  correspond to the  areas  of  the noise  level in a region to the
 left (AI) and to  the  right (Ar) of the GC peak of interest.
                                 8280 - A -  1
                                                          Revision
                                                         Date  September 1986

-------
     The  procedure  to  determine  the  S/N  1s as  follows:

     1.    Estimate  the average negative  peak  excursions  of the noise  (I.e.,
          the  low segment-E2-of   the  noise  envelope).    Line  £2  should pass
          through the  estimated  statistical  mean  of the negative-going noise
          excursions.   As  stated earlier,  1t  1s  Important to have the  signal
          offset  (0)   set  high  enough such  that  negative-going  noise  1s
     j     recorded„

     2.    Using the cross-hairs   of   the video  display terminal,  measure the
          peak area (A)  above a  baseline   corresponding to the mean negative
          noise value  (£2) and between the time  tj and  t2 where  the GC/MS peak
          intersects the baseline, £2.  Make note  of the time width t=t2-ti.

     3.    Following a  similar procedure   as  described  above, measure the area
          of the noise 1n a region to the   left (Aj) and to the right  (Ar)  of
          the GC/MS signal using a time  window twice the size of t, that Is,
          2 x t.

     The analyst must  sound  judgement  in   regard to  the proper  selection  of
Interference-free regions  1n  the  measurement   of  AI and  Ar.     It is not
recommended to perform these noise measurements   (Aj and Ar)  in  remote  regions
exceeding ten time widths (lot).

     4.   Compute the S/N using Eq.  1.

     NOTE:  If the noise does not occupy  at  least  10 percent  of the vertical
            axis (i.e., the noise envelope cannot be  defined  accurately),  then
            1t is  necessary to  amplify  the  vertical   axis  so that the noise
            occupies 20 percent of the terminal  display (see Figure 3).
                                8280 - A - 2
                                                         Revision
                                                         Date  September  1986

-------
                               FIGURE CAPTIONS

Figure 1.  Manual determination of S/N.
           The peak height (S) is measured between the mean noise (lines C and
           D).  These mean  signal  values  are  obtained  by tracing the line
           between the baseline average noise extremes, E\ and £21  and between
           the apex average noise  extremes,  £3  and  £4,  at the apex of the
           signal.  Note,  it  is  imperative  that the instrument's interface
           amplifier electronic's zero  offset  be  set  high enough such that
           negative-going baseline noise is recorded.

Figure 2.  Interactive determination of S/N.
           The peak area (A) is  measured  above the baseline average negative
           noise £2 and between times tj  and  t2-  The noise is obtained from
           the areas AI and Ar measured  to  the  left and to the right of the
           peak of interest using time windows Tj and Tr (Ti=Tr=2t).

Figure 3.  Interactive determination of S/N.
           A) Area measurements  without  amplification  of the vertical axis.
           Note that  the  noise  cannot  be  determined  accurately by visual
           means.   B)  Area  measurements  after  amplification  (10X) of the
           vertical axis so  that  the  noise  level occupies approximately 20
           percent of the display, thus enabling a better visual estimation of
           the baseline noise, Ej, £2, and C.
                                8280 - A - 3
                                                         Revision      0
                                                         Date  September 1986

-------
                                              O
                                              (O
                                              c.
                                              0>
                                              •M

                                              O





                                              c
                                              (O
                                               2!
                                               3
                                               D)
8280 - A -  4
                          Revision      0
                          Date   September 1986

-------
                                         t =
100 n
 90
 80
 70-
 60-
 50-
 40-
 30*
 20-
 10-
  0
                   = 558.10
   = 88.55
           i       i      r
         25:30 26:00  26:30
27:00   27:30 28:00
                       17 sec.
           Figure 2.  Interactive Determination of S/N.
                      8280 - A - 5
                                          Revision     p
                                          Date  September 1986

-------
100-

 90-

 80-

 70-

 60-

 50-

 40-

 30-

 20-

 10-

  0-
             A = 686.41
= 17.18
Ar= 13.32
        25:30  26:00  36:30  27:00  27:30  28:00
                         A = 706.59
                              Ar = 41.88
        26:30  26:00  26:30  27:00  27:30  28:00
      Figure 3.  Interactive Determination of S/N.
                   8280 - A - 6
                                         Revision     0
                                         Date  September 1986

-------
                                 APPENDIX B

        RECOMMENDED SAFETY AND HANDLING PROCEDURES FOR PCDD'S/PCDF'S


     1.  The human toxicology  of  PCDD/PCDF  is  not well  defined at present,
although the 2,3,7,8-TCDD isomer has been found to be acnegenic,  carcinogenic,
and teratogenic in the course of  laboratory animal studies.  The 2,3,7,8-TCDD
is a solid at room temperature, and  has a relatively low vapor pressure.  The
solubility of this compound in water is only about 200 parts-per-trillion,  but
the solubility in various organic  solvents  ranges from about 0.001 perent to
0.14 percent.  The physical properties  of  the 135 other tetra- through octa-
chlorinated PCDD/PCDF have not been  well established, although it is presumed
that the physical properties of these congeners are generally similar to those
of the 2,3,7,8-TCDD isomer.  On  the  basis of the available toxicological  and
physical property data for TCDD, this compound,  as well as the other PCDD and
PCDF, should be handled only  by  highly  trained personnel who are thoroughly
versed in the appropriate procedures, and who understand the associated risks.

     2.  PCDD/PCDF and samples  containing these are handled using essentially
the same techniques as  those  employed  1n handling radioactive or infectious
materials.  Well-ventilated, controlled-access  laboratories are required,  and
laboratory personel entering these laboratories should wear appropriate safety
clothing, including disposable coveralls,  shoe  covers,  gloves, and face and
head masks.  During analytical operations  which  may give rise to aerosols or
dusts,  personnel  should  wear  respirators  equipped  with  activated carbon
filters.  Eye protection equipment (preferably full face shields) must be worn
at all  times  while  working  in  the  analytical  laboratory with PCDD/PCDF.
Various  types  of  gloves  can  be  used  by  personnel,  depending  upon the
analytical operation being accomplished.  Latex gloves are generally utilized,
and when handling samples thought  to be particularly hazardous, an additional
set of gloves are also  worn  beneath  the  latex gloves (for example, Playtex
gloves supplied by American Scientific  Products, Cat. No. 67216).  Bench-tops
and other work surfaces  in  the  laboratory  should  be covered with plastic-
backed absorbent paper during all  analytical processing.  When finely divided
samples  (dusts,  soils, dry  chemicals)  are  processed,  removal of these from
sample  contaners,  as   well   as   other   operations,   including  weighing,
transferring, and mixing with  solvents,  should  all be accomplished within a
glove  box.   Glove boxes, hoods  and  the effluents  from mechanical vacuum pumps
and gas chromatographs  on  the  mass  spectrometers  should  be vented to the
atmosphere preferably only after passing  through HEPA particulate  filters and
vapor-sorbing charcoal.

      3.  All  laboratory  ware,  safety   clothing,   and other  items  potentially
contaminated with   PCDD/PCDF   in  the  course   of   analyses   must be carefully
secured and  subjected to proper  disposal.    When  feasible,  liquid wastes are
concentrated, and  the residues  are  placed  in  approved steel hazardous waste
drums  fitted with   heavy   gauge  polyethylene   liners.   Glass and  combustible
items  are  compacted using  a dedicated  trash compactor  used only  for hazardous
waste  materials  and then placed 1n   the   same  type  of disposal drum.  Disposal
of  accumulated  wastes   1s   periodically accomplished  by   high   temperature
incineration at  EPA-aproved facilities.


                                8280 - B  - 1
                                                         Revision      0
                                                         Date  September 1986

-------
     4.  Surfaces of laboratory benches,  apparatus and other appropriate areas
should be periodically subjected  to  surface  wipe tests using solvent-wetted
filter paper which Is then  analyzed  to   check for PCDD/PCDF contamination 1n
the laboratory.  Typically, if the detectable  level of TCDD or TCDF from such
a test is greater than 50  ng/m2,  this indicates the need for decontamination
of the laboratory.  A typical  action  limit 1n terms of surface contamination
of the other PCDD/PCDF (summed) is 500 ng/m2.   In the event of a spill  within
the laboratory, absorbent paper is  used  to  wipe up the spilled material and
this is then placed into a hazardous  waste drum.  The contaminated surface is
subsequently  cleaned  thoroughly   by   washing   with  appropriate  solvents
(methylene chloride followed by methanol)  and laboratory detergents.  This is
repeated until wipe tests  Indicate  that  the levels of surface contamination
are below the limits cited.

     5.  In  the  unlikely  event  that  analytical  personnel experience skin
contact with PCDD/PCDF   or  samples  containing  these, the contaminated skin
area should Immediately  be  thoroughly  scurbbed  using  mild soap and water.
Personnel involved in any such  accident  should  subsequently be taken to the
nearest medical facility, preferably  a  facility whose staff is knowledgeable
in  the  toxicology  of   chlorinated   hydrocarbons.     Again,  disposal  of
contaminated clothing is accomplished by placing it in hazardous waste drums.

     6.    It  is  desirable  that  personnel  working  in  laboratories where
PCDD/PCDF are  handled  be  given  periodic  physical  examinations  (at least
yearly).  Such examinations  should  Include  specialized tests, such as those
for urinary porphyrlns  and  for  certain  blood  parameters which, based upon
published clinical  observations,  are  appropriate  for  persons  who  may be
exposed to PCDD/PCDF.  Periodic  facial  photographs  to document the onset of
dermatologic problems are also advisable.
                                 8280 - B - 2
                                                          Revision      0
                                                          Date  September 1986

-------
                                                                   Page  1  of  2



                   DIOXIN SAMPLE DATA SUMMARY FORM 8280-1



LAB NAME 	     CONTRACT No.	


CASE No. 	

                                          QUANTITY FOUND (ng/g)


SAMPLE NO.     FILE NAME       TCDD      PeCDD      HxCDD      HpCDD      OCDD
 DATA  RELEASE  AUTHORIZED  BY
                                 8280 - B -  3
                                                          Revision       0
                                                          Date   September  1986

-------
                                                                   Page  2  of 2



                   DIOXIN SAMPLE DATA SUMMARY FORM 8280-1



LAB NAME	CONTRACT No. 	


CASE No. 	

                                          QUANTITY FOUND (ng/g)


SAMPLE NO.     FILE NAME       TCDF      PeCDF      HxCDF      HpCDF      OCDF
                                 8280 - B - 4
                                                          Revision
                                                          Date  September 1986

-------
                                                                   Page  1  of  2



                  DIOXIN SAMPLE DATA SUMMARY FORM 8280-1-W



LAB NAME 	      CONTRACT No.  	


CASE No. 	

                                          QUANTITY FOUND (ug/L)


SAMPLE NO.     FILE NAME       TCDD      PeCDD      HxCDD      HpCDD      OCDD
 DATA  RELEASE AUTHORIZED  BY
                                 8280 -  B - 5
                                                         Revision
                                                         Date  September 1986

-------
                                                                   Page 2 of 2



                  DIOXIN SAMPLE DATA SUMMARY FORM 8280-1-W



LAB NAME	CONTRACT No. 	


CASE No. 	

                                          QUANTITY FOUND (ug/L)


SAMPLE NO.     FILE NAME       TCDF      PeCDF      HxCDF      HpCDF      OCDF
                                 8280 - B - 6
                                                          Revision
                                                          Date  September 1986

-------
                     DIOXIN RAW SAMPLE DATA FORM 8280-2
LAB NAME 	  ANALYST(s)  	  CASE No.
SAMPLE No.                   TYPE OF SAMPLE	CONTRACT No.
SAMPLE SIZE           % MOISTURE             FINAL EXTRACT VOLUME
EXTRACTION METHOD 	ALIQUOT USED FOR ANALYSIS


CLEAN UP OPTION
CONCENTRATION FACTOR                  DILUTION FACTOR
DATE EXTRACTED                        DATA ANALYZED
VOLUME 13Ci2-l,2,3,4-TCDD ADDED 	 TO SAMPLE VOLUME
VOLUME INJECTED 	 Wt 13Ci2-l,2,3,4-TCDD ADDED
Wt 13Ci2-2,3,7,8-TCDD ADDED 	 13Ci2-2,3,7,8-TCDD % RECOVERY


Wt 13Ci2-2,3,7,8-OCDD ADDED 	 13C12-OCDD % RECOVERY 	
13Ci2-2,3,7,8-TCDD RRF 	  13Ci2-OCDD RRF 	

                             13Ci2-2,3,7,8-TCDD

AREA 332 	 AREA 334 	 RATIO 332/334 _


13Ci2-OCDD AREA 470 	  AREA 472 	  RATIO 470/472


RT 2,3,7,8-TCDD (Standard) 	  RT 2,3,7,8-TCDD (Sample) 	


13C12-2,3,7,8-TCDD -  13Ci2-l,2,3,4-TCDD Percent Valley 	
                                8280 - B - 7
                                                         Revision      0
                                                         Date  September 1986

-------
              DIOXIN INITIAL CALIBRATION STANDARD DATA SUMMARY

                                 FORM 8280-3

                                          CASE No.
Lab Name
Date of Initial Calibration

Relative to
                        Contract No.

                        Analyst(s)
CALIBRATION
STANDARD
RRF
 1
                           RRF     RRF   RRF
                            2       34
                                              or 13Ci2-l,2r3,4-TCDD_
RRF
 5
MEAN    %RSD
TCDD
PeCDD
HxCDD
HpCDD
OCDD
TCDF
 PeCDF
 HxCDF
 HpCDF
 OCDF
                                 8280 - B - 8
                                                          Revision      0
                                                          Date  September 1986

-------
                           FORM 8280-3 (Continued)


                           CONCENTRATIONS IN PG/UL

                  1        2345
TCDD
PeCDD
HxCDD
HpCDD
OCDD
TCDF
PeCDF
HxCDF
HpCDF
OCDF
                                 8280 - B - 9
                                                          Revision
                                                          Date  September 1986

-------
                    DIOXIN CONTINUING CALIBRATION SUMMARY

                                 FORM 8280-4



                                         CASE No.
Lab Name
Date of Initial Calibration
Relative to 13Ci2-2,3,7,8-TCDD_
                     Contract  No.

                     Analyst(s)	
                     or 13C12-1,2,3,4-TCDD
COMPOUND
RRF
RRF
%D
TCDD
PeCDD
HxCDD
HpCDD
OCDD
TCDF
 PeCDF
HxCDF
 HpCDF
 OCDF
                                 8280 - B - 10
                                                          Revision      0
                                                          Date  September 1986

-------
                    DIOXIN RAW SAMPLE DATA FORM 8280-5-A



LAB NAME 	  ANALYST(s) 	  CASE No.


CONTRACT No.                                SAMPLE No. 	
TCDD REQUIRED 320/322 RATIO WINDOW IS 0.65 - 0.89
QUANTITATED FROM 2,3,7,8-TCDD 	  1,2,3,4-TCDD 	  RRF
SCAN I  RRT   AREA       AREA       AREA       320/      CONFIRM
              322        320        257        322       AS TCDD
                                                           Y/N      CONC,
                                             TOTAL  TCDD
 TCDF  REQUIRED 304/306  RATIO WINDOW IS  0.65  -  0.89

 QUANTITATED FROM 2,3,7,8-TCDD	 1,2,3,4-TCDD 	   RRF
 SCAN  #   RRT   AREA       AREA       AREA       304/       CONFIRM
               306        304        243         306        AS  TCDD
 _^_^_	Y/N       CONC,
                                             TOTAL  TCDD
                                 8280 - B - 11
                                                          Revision
                                                          Date   September  1986

-------
                    DIOXIN RAW SAMPLE DATA FORM 8280-5-B
LAB NAME
                  ANALYST(s)
                                      CASE No.
CONTRACT No.
                                   SAMPLE No.
PeCDD REQUIRED 320/322 RATIO WINDOW IS 0.55 - 0.75
QUANTITATED FROM 2,3,7,8-TCDD

SCAN #  RRT   AREA      AREA
              356       358
                                 1,2,3,4-TCDD
                         AREA
                         354
                             AREA
                             293
                           3587
                           356
                                                  RRF
                           CONFIRM
                           AS PeCDD
                             Y/N
                                                                         CONC.
                                           TOTAL  PeCDD
 PeCDF  REQUIRED 342/340  RATIO WINDOW IS  0.55  -  0.75

 QUANTITATED FROM 2,3,7,8-TCDD 	  1,2,3,4-TCDD

 SCAN
RRT
AREA
340
AREA
342
AREA
338
AREA
277
342/
340
                                                          RRF
CONFIRM
AS PeCDF
  Y/N
                                                                          CONC.
                                            TOTAL PeCDF
                                 8280 - B - 12
                                                          Revision      0
                                                          Date  September 1986

-------
                    DIOXIN RAW SAMPLE DATA FORM 8280-5-C
LAB NAME
ANALYST(s)
         CASE No.
CONTRACT No.
                  SAMPLE No.
HxCDD REQUIRED 392/390 RATIO WINDOW IS 0.69 - 0.93
QUANTITATED FROM 2,3,7,8-TCDD

SCAN #  RRT   AREA      AREA
              390       392
                1,2,3,4-TCDD
       AREA
       388
AREA
327
392/
390
                     RRF
CONFIRM
AS HxCDD
  Y/N
                                                                         CONC.
                                           TOTAL HxCDD

HxCDF REQUIRED 376/374
QUANTITATED FROM 2,3,7
SCAN # RRT AREA
376
RATIO WINDOW IS
,8-TCDD
AREA
374
AREA
372
0.69 - 0.93
1,2,3, 4-TCDD
AREA
311
376/
374
RRF
CONFIRM
AS HxCDF
Y/N CONC.
                                          TOTAL HxCDF
                                8280 -  B - 13
                                                         Revision      0
                                                         Date  September 1986

-------
                    DIOXIN RAW SAMPLE DATA FORM 8280-5-D
LAB NAME
ANALYST(s)
         CASE No.
CONTRACT No.
                  SAMPLE No.
HpCDD REQUIRED 426/444 RATIO WINDOW IS 0.83 - 1.12
QUANTITATED FROM 2,3,7,8-TCDD

SCAN I  RRT   AREA      AREA
              424       426
                1,2,3,4-TCDD
       AREA
       422
AREA
361
4267
424
                     RRF
CONFIRM
AS HpCDD
  Y/N
                                                                         CONC.
                                           TOTAL HpCDD

HpCDF REQUIRED 410/408
QUANTITATED FROM 2,3,7
SCAN # RRT AREA
408
RATIO WINDOW IS
,8-TCDD
AREA AREA
410 406
0.83 - 1.12
1,2,3,4-TCDD
AREA
345
410/
408
RRF
CONFIRM
AS HpCDF
Y/N CONC.
                                           TOTAL HpCDF
                                 8280 - B - 14
                                                          Revision      0
                                                          Date  September 1986

-------
                    DIOXIN RAW SAMPLE DATA FORM 8280-5-E



LAB NAME 	  ANALYST(s) 	  CASE No..

CONTRACT No.                                  SAMPLE No. 	
OCDD REQUIRED 458/460 RATIO WINDOW IS 0.75 - 1.01
QUANTITATED FROM 2,3,7,8-TCDD 	    1,2,3,4-TCDD 	  RRF
SCAN I  RRT   AREA        AREA      AREA       458/      CONFIRM
              460         458       395        460       AS OCDD
                             	Y/N      CONC.
                                           TOTAL OCDD
OCDF REQUIRED 442/444 RATIO WINDOW IS 0.75 - 1.01

QUANTITATED FROM 2,3,7,8-TCDD 	  1,2,3,4-TCDD 	  RRF
SCAN #  RRT   AREA        AREA      AREA       442/      CONFIRM
              444         442       379        444       AS OCDF
	Y/N	CONC.
                                           TOTAL OCDF
                                 8280 - B -  15
                                                          Revision
                                                          Date   September  1986

-------
            DIOXIN SYSTEM PERFORMANCE CHECK ANALYSIS FORM 8280-6
LAB NAME
                    CASE  No.
BEGINNING DATE
ENDING DATE
       TIME
       TIME
                   CONTRACT No..
                    ANALYST(s)
PC SOLUTION IDENTIFIER
PCDD's
                       ISOTOPIC RATIO CRITERIA MEASUREMENT
  IONS
RATIOED
RATIO AT
BEGINNING OF
12 HOUR PERIOD
RATIO AT
END OF 12   ACCEPTABLE
HOUR PERIOD   WINDOW
Tetra
320/322
                               0.65-0.89
Penta
358/356
                               0.55-0.75
Hexa
392/390
                               0.69-0.93
Hepta
426/424
                               0.83-1.12
Octa
458/460
                               0.75-1.01
 PCDF's
 Tetra
304/306
                               0.65-0.89
 Penta
342-340
                               0.55-0.75
 Hexa
376-374
                               0.69-0.93
 Hepta
410/408
                               0.83-1.12
 Octa
442/444
                               0.75-1.01
 Ratios out of criteria
 PCDD
 PCDF
       Beginning
      _ out  of
        out  of
                             End
                            out of
                            out of
 NOTE:  One form is required for each 12 hour period samples are analyzed.
                                 8280 - B - 16
                                                          Revision      0
                                                          Date  September 1986

-------
                                  METHOD 8290

   POLYCHLORINATED DIBENZODIOXINS  (PCDDs) AND POLYCHLORINATED DIBENZOFURANS
          (PCDFs)BY HIGH-RESOLUTION GAS CHROMATOGRAPHY/HIGH-RESOLUTION
                         MASS SPECTROMETRY (HRGC/HRMS)
1.0   SCOPE AND APPLICATION

      1.1   This method provides procedures for the detection and quantitative
measurement of polychlorinated dibenzo-p-dioxins (tetra- through octachlorinated
homologues;   PCDDs),   and   polychlorinated  dibenzofurans   (tetra-  through
octachlorinated homologues; PCDFs)  in  a variety of environmental matrices  and at
part-per-trillion  (ppt)   to  part-per-quadrillion  (ppq)  concentrations.   The
following compounds can be determined by this method:
                        Compound Name
CAS No"
2,3,7,8-Tetrachlorodibenzo-p-dioxin (TCDD)
1,2,3,7 , 8-Pentachl orodi benzo-p-di oxi n (PeCDD)
1,2,3,6,7,8-Hexachlorodibenzo-p-dioxin (HxCDD)
1,2, 3, 4, 7, 8-Hexachl orodi benzo-p-di oxi n (HxCDD)
1,2, 3, 7, 8, 9-Hexachl orodi benzo-p-di oxi n (HxCDD)
1,2, 3, 4, 6, 7, 8-Heptachl orodi benzo-p-di oxi n (HpCDD)
1,2,3,4,6,7,8,9-Octachlorodibenzo-p-dioxin (OCDD)
2,3,7,8-Tetrachlorodibenzofuran (TCDF)
1,2,3,7,8-Pentachlorodibenzofuran (PeCDF)
2,3,4,7,8-Pentachlorodibenzofuran (PeCDF)
1,2,3,6,7, 8-Hexachl orodi benzof uran (HxCDF )
1,2, 3, 7, 8, 9-Hexachl orodi benzofuran (HxCDF)
1,2,3,4,7 , 8-Hexachl orodi benzof uran (HxCDF)
2, 3, 4, 6, 7, 8-Hexachl orodi benzof uran (HxCDF)
1,2,3,4,6,7 , 8-Heptachl orod i benzofuran ( HpCDF)
1,2,3,4,7,8 , 9-Heptachl orod i benzofuran (HpCDF)
1 , 2 , 3 , 4 , 6 , 7 , 8 , 9 -Octachl orod i benzofuran (OCDF)
1746-01-6
40321-76-4
57653-85-7
39227-28-6
19408-74-3
35822-39-4
3268-87-9
51207-31-9
57117-41-6
57117-31-4
57117-44-9
72918-21-9
70648-26-9
60851-34-5
67562-39-4
55673-89-7
39001-02-0
      a     Chemical Abstract Service Registry Number

      1.2   The  analytical  method calls  for the  use of  high-resolution  gas
chromatography and  high-resolution  mass spectrometry  (HRGC/HRMS)  on purified
sample  extracts.    Table  1  lists the  various  sample types  covered  by  this
analytical protocol, the 2,3,7,8-TCDD-based method calibration limits (MCLs), and
other  pertinent  information.   Samples  containing concentrations  of specific
congeneric analytes (PCDDs and PCDFs) considered within the scope of this method
that are greater than ten times the  upper MCLs  must  be analyzed  by a protocol
designed for such concentration levels,  e.g., Method 8280.  An optional method
for reporting the analytical results using a 2,3,7,8-TCDD toxicity equivalency
factor (TEF) is described.
                                   8290 - 1
       Revision  0
   September 1994

-------
      1.3   The sensitivity of this method is dependent upon the level of inter-
ferences within a given matrix.  The calibration range of the method for a I L
water sample is 10  to 2000 ppq for TCDD/TCDF and PeCDD/PeCDF,  and  1.0 to 200 ppt
for a  10 g soil,  sediment,  fly ash, or tissue  sample for  the  same  analytes
(Table 1).  Analysis of a one-tenth aliquot of the sample permits measurement of
concentrations up to 10 times  the upper MCL.  The  actual limits of detection and
quantitation will differ from the  lower  MCL,  depending on the complexity of the
matrix.

      1.4   This method is designed for use by analysts who  are experienced with
residue analysis and skilled in HRGC/HRMS.

      1.5   Because of the  extreme toxicity  of  many of these  compounds,  the
analyst must  take  the  necessary precautions to  prevent exposure to materials
known or believed to contain  PCDDs or PCDFs.   It is the  responsibility of the
laboratory personnel to ensure that safe handling procedures are employed.  Sec.
11 of this method discusses safety procedures.
2.0   SUMMARY OF METHOD
                                                                    e specific
      2.1   This procedure  uses  matrix specific extraction,  analyt
cleanup, and HRGC/HRMS analysis techniques.

      2.2   If  interferences  are  encountered,  the  method  provides  selected
cleanup  procedures  to aid  the analyst  in their  elimination.    A  simplified
analysis flow chart is presented at the end of this method.

      2.3   A specified amount (see Table  1) of soil, sediment, fly ash, water,
sludge (including paper pulp), still bottom, fuel oil, chemical reactor residue,
fish  tissue,  or human  adipose tissue  is spiked  with  a  solution  containing
specified amounts of  each of  the  nine  isotopically (13C12) labeled PCDDs/PCDFs
listed  in Column  1  of Table  2.   The  sample is then extracted  according  to a
matrix specific extraction procedure. Aqueous samples that are judged to contain
1 percent or  more  solids,  and solid samples  that  show an  aqueous  phase,  are
filtered, the solid  phase (including the filter) and the aqueous phase extracted
separately,  and the extracts  combined  before extract  cleanup.  The extraction
procedures are:

      a)    Toluene:  Soxhlet extraction for soil,  sediment,  fly ash, and paper
            pulp samples;

      b)    Methylene chloride:  liquid-liquid extraction for water samples;

      c)    Toluene:   Dean-Stark  extraction for fuel oil,  and  aqueous sludge
            samples;

      d)    Toluene extraction for still bottom samples;

      e)    Hexane/methylene  chloride:     Soxhlet  extraction   or  methylene
            chloride:  Soxhlet extraction for fish tissue samples; and

      f)    Methylene chloride extraction for human adipose tissue samples.
                                   8290 - 2
                                                                    Revision 0
                                                                September 1994

-------
      g)    As an option, all solid samples (wet or dry) may be extracted with
            toluene using a Soxhlet/Dean Stark extraction system.

      The decision  for the selection of an extraction  procedure for chemical
reactor residue samples  is based on the appearance (consistency, viscosity) of
the samples.  Generally,  they  can be handled according to the procedure used for
still bottom  (or chemical sludge) samples.

      2.4   The extracts are  submitted to an  acid-base washing treatment and
dried.  Following  a  solvent exchange step, the extracts are cleaned  up by column
chromatography on alumina, silica gel, and activated carbon.

            2.4.1 The  extracts  from adipose  tissue  samples are  treated with
      silica gel  impregnated with sulfuric  acid before chromatography on acidic
      silica gel,  neutral alumina, and activated carbon.

            2.4.2 Fish tissue and paper pulp extracts are subjected to an acid
      wash  treatment  only,  prior to  chromatography  on  alumina and activated
      carbon.

      2.5   The preparation  of  the  final  extract  for HRGC/HRMS  analysis  is
accomplished  by  adding  10  to  50 /xL  (depending  on  the  matrix)  of  a  nonane
solution containing  50 pg/juL  of the recovery  standards 13C12-1,2,3,4-TCDD and
13C12-l,2,3,7,8,9-HxCDD (Table 2).  The former  is used to determine the percent
recoveries of tetra- and pentachlorinated  PCDD/PCDF congeners, while the latter
is  used  to  determine  the   percent   recoveries  of  the  hexa-,  hepta-  and
octachlorinated PCDD/PCDF congeners.

      2.6   Two juL  of  the  concentrated  extract are  injected into an HRGC/HRMS
system capable of performing selected ion monitoring at resolving powers of at
least 10,000  (10 percent valley definition).

      2.7   The  identification  of  OCDD  and  nine   of  the  fifteen  2,3,7,8-
substituted congeners  (Table 3), for which a 13C-labeled standard is available
in the sample fortification and recovery standard  solutions  (Table 2), is based
on their elution at their exact retention time (within 0.005 retention time units
measured in the routine  calibration) and the simultaneous detection of the two
most abundant  ions  in  the molecular  ion  region.   The  remaining  six  2,3,7,8-
substituted congeners  (i.e.,  2,3,4,7,8-PeCDF;  1,2,3,4,7,8-HxCDD;  1,2,3,6,7,8-
HxCDF; 1,2,3,7,8,9-HxCDF; 2,3,4,6,7,8-HxCDF, and 1,2,3,4,7,8,9-HpCDF), for which
no carbon-labeled internal standards are available in the sample fortification
solution, and all  other  PCDD/PCDF congeners are identified when their relative
retention times fall within their respective PCDD/PCDF retention time windows,
as established from  the routine calibration data,  and the simultaneous detection
of the two most abundant ions in the molecular ion region.  The identification
of OCDF is based on its  retention time relative to 13C12-OCDD and the simultaneous
detection  of  the  two  most   abundant  ions   in  the   molecular  ion  region.
Identification also  is  based on  a comparison of the ratios  of the integrated ion
abundance of the molecular ion species to their theoretical abundance ratios.

      2.8   Quantitation of the individual  congeners, total PCDDs and total PCDFs
is achieved in conjunction  with  the  establishment of a multipoint (five points)
                                   8290 - 3                         Revision 0
                                                                September 1994

-------
calibration curve for each  homologue, during which each calibration solution is
analyzed once.


3.0   INTERFERENCES

      3.1   Solvents, reagents, glassware and other sample processing hardware
may yield  discrete artifacts or  elevated  baselines  that may  cause  misinter-
pretation of the chromatographic data  (see references  1  and  2.)   All  of these
materials must be demonstrated to be free from interferants under the conditions
of analysis by performing laboratory method blanks.  Analysts  should avoid using
PVC gloves.

      3.2   The  use  of  high  purity  reagents  and  solvents  helps  minimize
interference problems.  Purification of  solvents by  distillation  in  all-glass
systems may be necessary.

      3.3   Interferants coextracted from the sample will  vary considerably from
matrix to matrix.  PCDDs and PCDFs are often  associated with  other interfering
chlorinated substances such as polychlorinated biphenyls (PCBs), polychlorinated
diphenyl  ethers  (PCDPEs),   polychlorinated  naphthalenes, and  polychlorinated
alkyldibenzofurans,  that  may  be  found  at  concentrations  several  orders  of
magnitude  higher than the  analytes  of  interest.   Retention times  of  target
analytes  must be  verified  using  reference  standards.   These  values  must
correspond to  the  retention  time  windows established in Sec.  8.1.1.3.   While
cleanup  techniques are provided  as  part of  this  method,  unique  samples  may
require additional  cleanup steps to achieve lower detection  limits.

      3.4   A  high-resolution capillary  column  (60 m DB-5, J&W Scientific,  or
equivalent) is used in this method.   However, no  single  column is  known  to
resolve all isomers.  The 60 m  DB-5 GC  column  is capable of 2,3,7,8-TCDD isomer
specificity  (Sec.  8.1.1).    In order  to determine  the  concentration  of  the
2,3,7,8-TCDF  (if detected   on  the DB-5  column),  the  sample extract must  be
reanalyzed on  a column capable of 2,3,7,8-TCDF isomer  specificity (e.g., DB-225,
SP-2330, SP-2331, or equivalent).


4.0   APPARATUS AND MATERIALS

      4.1   High-Resolution    Gas    Chromatograph/High-Resolution    Mass
Spectrometer/Data System (HRGC/HRMS/DS) - The  GC must be  equipped for temperature
programming, and all required accessories must be available,  such as  syringes,
gases, and capillary columns.

            4.1.1 GC Injection  Port  - The GC injection port must be designed for
      capillary  columns.     The  use  of splitless  injection  techniques  is
      recommended.   On column  1  juL  injections can  be  used  on the 60  m DB-5
      column.   The use of  a moving needle injection port  is also acceptable.
      When using the method described in  this  protocol, a 2 juL injection volume
      is  used consistently  (i.e.,  the  injection  volumes  for all  extracts,
      blanks,  calibration solutions and the performance  check  samples are 2 /LtL).
      One  ^L  injections  are  allowed;  however,   laboratories  must  remain
                                   8290 - 4                         Revision 0
                                                                September 1994

-------
consistent throughout the analyses by using the same injection volume at
all times.

      4.1.2 Gas Chromatograph/Mass Spectrometer  (GC/MS)  Interface - The
GC/MS interface components should withstand 350°C.  The interface must be
designed  so  that  the separation  of 2,3,7,8-TCDD  from  the  other TCDD
isomers  achieved  in the  gas  chromatographic  column is  not appreciably
degraded.  Cold spots or active surfaces (adsorption sites) in the GC/MS
interface can cause peak tailing and peak broadening.  It is recommended
that  the GC  column  be  fitted directly  into  the mass  spectrometer ion
source without  being exposed to  the ionizing electron  beam.   Graphite
ferrules should be avoided in the injection port because they may adsorb
the PCDDs and PCDFs.  Vespel™,  or equivalent, ferrules are recommended.

      4.1.3 Mass  Spectrometer  -  The   static resolving  power  of  the
instrument must be maintained  at a minimum of 10,000  (10 percent valley).

      4.1.4 Data System - A dedicated data system is employed to control
the  rapid multiple-ion  monitoring  process  and  to  acquire the  data.
Quantitation data (peak areas  or peak heights)  and SIM traces  (displays of
intensities of each ion signal being monitored  including the lock-mass ion
as a  function of  time) must  be  acquired during the  analyses and stored.
Quantitations may be reported based upon computer generated peak areas or
upon  measured peak heights  (chart recording).  The  data  system must be
capable of acquiring data at  a minimum  of 10 ions  in  a single scan. It is
also recommended to have a data system capable of switching to different
sets  of  ions   (descriptors)  at  specified times  during  an  HRGC/HRMS
acquisition.   The  data  system should be able  to  provide  hard copies of
individual  ion  chromatograms  for  selected   gas  chromatographic  time
intervals.  It should also be  able to acquire  mass spectral peak profiles
(Sec. 8.1.2.3)  and provide hard copies of peak  profiles to demonstrate the
required resolving power.  The data system should permit the measurement
of noise on the base line.

      NOTE: The detector  ADC zero setting must  allow  peak-to-peak measure-
            ment of the  noise on the  base line of every monitored channel
            and  allow for good  estimation of the  instrument resolving
            power.  In Figure 2,  the  effect of different zero settings on
            the measured resolving power is shown.

4.2   GC Columns

      4.2.1 In order to have an isomer specific determination for 2,3,7,8-
TCDD  and  to  allow the detection  of  OCDD/OCDF within a  reasonable time
interval  in  one HRGC/HRMS analysis,  use of the  60  m DB-5  fused  silica
capillary column  is recommended.  Minimum acceptance criteria must  be
demonstrated and  documented  (Sec.  8.2.2).   At the  beginning  of each 12
hour  period  (after mass  resolution and  GC  resolution  are demonstrated)
during which sample extracts  or concentration  calibration solutions will
be analyzed,  column operating  conditions must be attained  for the required
separation on the  column  to  be  used for samples.   Operating conditions
known to produce acceptable results with the recommended column are shown
in Sec. 7.6.
                             8290 - 5                         Revision 0
                                                          September 1994

-------
            4.2.2 Isomer specificity  for all  2,3,7,8-substituted  PCDDs/PCDFs
      cannot  be  achieved  on  the  60  m  DB-5  GC  column  alone.   In  order  to
      determine the proper  concentrations of the individual 2,3,7,8-substituted
      congeners,  the sample extract must be reanalyzed on  another GC column that
      resolves the isomers.

            4.2.3 30 m DB-225 fused silica capillary column,  (J&W Scientific) or
      equivalent.

      4.3   Miscellaneous Equipment and Materials  -  The following list of items
does not necessarily constitute an exhaustive compendium of the equipment needed
for this analytical method.

            4.3.1 Nitrogen evaporation apparatus with variable flow rate.

            4.3.2 Balances  capable  of  accurately   weighing  to  0.01  g  and
      0.0001 g.

            4.3.3 Centrifuge.

            4.3.4 Water bath,  equipped with concentric ring covers and capable
      of being temperature controlled within + 2°C.

            4.3.5 Stainless steel  or  glass  container  large  enough to  hold
      contents of one pint sample containers.

            4.3.6 Glove box.

            4.3.7 Drying oven.

            4.3.8 Stainless steel spoons and spatulas.

            4.3.9 Laboratory hoods,

            4.3.10      Pipets, disposable, Pasteur, 150 mm long x 5 mm ID.

            4.3.11      Pipets,  disposable,   serological,   10  ml,   for  the
      preparation of the carbon columns specified in Sec. 7.5.3.

            4.3.12      Reaction vial,  2 ml, silanized amber glass (Reacti-vial,
      or equivalent).

            4.3.13      Stainless steel meat grinder with a  3 to 5 mm hole size
      inner plate.

            4.3.14      Separatory funnels, 125 ml and 2000 ml.

            4.3.15      Kuderna-Danish concentrator, 500 ml, fitted with 10 ml
      concentrator tube and three ball Snyder column.

            4.3.16      Teflon™ or carborundum (silicon  carbide) boiling chips
      (or equivalent), washed with hexane before use.
                                   8290 - 6                         Revision 0
                                                                September 1994

-------
      NOTE: Teflon™ boiling chips may float  in  methylene  chloride,  may
            not  work  in  the  presence of  any water  phase,  and  may  be
            penetrated by nonpolar organic compounds.

      4.3.17      Chromatographic columns,  glass, 300 mm x 10.5 mm, fitted
with Teflon™ stopcock.

      4.3.18      Adapters for concentrator tubes.

      4.3.19      Glass   fiber   filters,   0.70   urn,  Whatman   GFF,   or
equivalent.

      4.3.20      Dean-Stark trap, 5  or  10 ml, with T-joints,  condenser
and 125 ml flask.

      4.3.21      Continuous liquid-liquid extractor.

      4.3.22      All  glass Soxhlet apparatus, 500 ml flask.

      4.3.23      Soxhlet/Dean Stark extractor (optional), all glass,  500
ml flask.

      4.3.24      Glass funnels, sized to hold 170 ml of liquid.

      4.3.25      Desiccator.

      4.3.26      Solvent reservoir  (125 ml), Kontes;  12.35  cm diameter
(special order item),  compatible with gravity carbon column.

      4.3.27      Rotary  evaporator with a temperature  controlled water
bath.

      4.3.28      High  speed  tissue  homogenizer, equipped with  an  EN-8
probe, or equivalent.

      4.3.29      Glass wool,  extracted with methylene chloride, dried and
stored in a clean glass jar.

      4.3.30      Extraction jars, glass,  250  ml, with teflon lined screw
cap.

      4.3.31      Volumetric flasks,  Class A - 10 ml to 1000  ml.

      4.3.32      Glass vials, 1 dram (or metric  equivalent).

      NOTE: Reuse of glassware should be minimized  to  avoid  the risk of
            contamination.    All  glassware  that  is  reused   must   be
            scrupulously cleaned as soon  as possible after use,  according
            to the  following  procedure:   Rinse  glassware with  the  last
            solvent used in  it.  Wash with hot  detergent water, then rinse
            with copious  amounts of  tap water  and several  portions  of
            organic-free  reagent water.  Rinse with  high purity acetone
                             8290 - 7                         Revision 0
                                                          September 1994

-------
                  and hexane and store it inverted or capped with solvent rinsed
                  aluminum foil in a clean environment.
5.0   REAGENTS AND STANDARD SOLUTIONS

      5.1   Organic-free reagent water - All references to water in this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.2   Column Chromatography Reagents

            5.2.1 Alumina,   neutral,   80/200   mesh   (Super   1,   Woelm®,   or
      equivalent).   Store in  a sealed container at room  temperature,  in  a
      desiccator, over self-indicating silica  gel.

            5.2.2 Alumina, acidic AG4,  (Bio Rad Laboratories catalog #132-1240,
      or equivalent).  Soxhlet extract with methylene chloride for 24 hours if
      blanks show contamination,  and activate by heating  in a foil covered glass
      container for 24 hours at  190°C.  Store in  a glass  bottle  sealed with  a
      Teflon™ lined screw cap.

            5.2.3 Silica gel, high  purity  grade,  type 60,  70-230 mesh; Soxhlet
      extract with methylene  chloride for 24 hours if blanks show contamination,
      and activate by heating in a  foil covered glass container for 24 hours at
      190°C.   Store in a  glass  bottle  sealed with a Teflon™ lined screw cap.

            5.2.4 Silica gel  impregnated with sodium  hydroxide.   Add one part
      (by weight)  of 1 M  NaOH  solution  to two  parts (by weight)  silica gel
      (extracted and activated) in a screw cap bottle and mix with a gl
      until  free of lumps.  Store in a glass bottle  sealed with a Teflon1
.ass rod
™ lined
      screw cap.
            5.2.5 Silica gel impregnated with 40  percent  (by  weight)  sulfuric
      acid.  Add  two parts (by weight) concentrated sulfuric acid to three parts
      (by weight) silica gel  (extracted  and activated),  mix with  a glass  rod
      until free of lumps,  and store in a screw capped glass  bottle.   Store in
      a glass bottle sealed with a Teflon™ lined  screw cap.

            5.2.6 Celite 545® (Supelco),  or equivalent.

            5.2.7 Active carbon  AX-21  (Anderson Development Co., Adrian, MI), or
      equivalent, prewashed with methanol and dried in vacuo at 110°C.  Store in
      a glass bottle sealed with a Teflon™ lined  screw cap.

      5.3   Reagents

            5.3.1 Sulfuric acid, H2S04, concentrated, ACS grade, specific gravity
      1.84.

            5.3.2 Potassium  hydroxide,  KOH, ACS  grade,   20  percent  (w/v)  in
      organic-free reagent water.
                                   8290 - 8                         Revision 0
                                                                September 1994

-------
            5.3.3 Sodium chloride, NaCl,  analytical  reagent,  5  percent  (w/v) in
      organic-free reagent water.

            5.3.4 Potassium carbonate, K2C03,  anhydrous,  analytical  reagent.

      5.4   Desiccating agent

            5.4.1 Sodium sulfate (powder, anhydrous), Na2S04. Purify by heating
      at 400°C  for  4 hours  in  a shallow  tray,  or by precleaning  the sodium
      sulfate with methylene chloride.  If the sodium sulfate  is precleaned with
      methylene chloride,  a  method  blank must be analyzed, demonstrating that
      there is no interference from the  sodium sulfate.

      5.5   Solvents

            5.5.1 Methylene chloride, CH2C12.   High purity, distilled in glass
      or highest available purity.

            5.5.2 Hexane,  CeH14.   High purity,  distilled  in glass  or highest
      available purity.

            5.5.3 Methanol, CH3OH.  High  purity,  distilled in glass or highest
      available purity.

            5.5.4 Nonane,  C9H20.   High purity,  distilled  in glass  or highest
      available purity.

            5.5.5 Toluene, C6H5CH3.  High purity, distilled in glass or highest
      available purity.
            5.5.6 Cyclohexane, C6H12,
      available purity.

            5.5.7 Acetone,
      available purity.
CH3COCH3.
           High purity, distilled in glass or highest
High purity,  distilled  in glass or highest
      5.6   High-Resolution Concentration Calibration Solutions (Table 5)  - Five
nonane solutions containing  unlabeled (totaling 17)  and carbon-labeled (totaling
11) PCDDs and PCDFs at known  concentrations are used to calibrate the instrument.
The concentration ranges are homologue dependent, with the lowest  values for the
tetrachlorinated dioxin  and furan  (1.0 pg/^L) and the  highest  values  for the
octachlorinated congeners (1000 pg//iL).

            5.6.1 Depending  on  the  availability  of materials,  these  high-
      resolution concentration calibration  solutions may be obtained from the
      Environmental Monitoring Systems Laboratory,  U.S.  EPA, Cincinnati,  Ohio.
      However,  additional secondary standards must be obtained from commercial
      sources,  and solutions should be prepared in the analyst's laboratory.  It
      is the responsibility  of the  laboratory  to  ascertain that the calibration
      solutions  received   (or  prepared)  are  indeed   at  the  appropriate
      concentrations before they are used to analyze samples.
                                   8290 - 9
                                         Revision 0
                                     September 1994

-------
            5.6.2 Store  the  concentration  calibration  solutions  in  1  ml
      mini vials at room temperature in the dark.

      5.7   GC Column Performance Check  Solution  -  This solution contains the
first and last eluting  isomers  for  each  homologous  series from tetra- through
heptachlorinated congeners.  The solution also contains a series of other TCDD
isomers for  the  purpose of  documenting  the chromatographic  resolution.   The
13Cl2-2,3,7,8-TCDD is also present.   The laboratory is required to use nonane as
the solvent  and adjust the volume so  that the final  concentration does not exceed
100 pg/juL per congener.  Table 7 summarizes the  qualitative composition (minimum
requirement) of this performance evaluation solution.

      5.8   Sample Fortification Solution  -  This  nonane solution contains the
nine internal  standards  at the nominal concentrations that are listed in Table 2.
The solution contains at least one carbon-labeled standard for each homologous
series, and it is used to measure the concentrations of the native substances.
(Note that 13C12-OCDf: is not present in the solution.)

      5.9   Recovery Standard  Solution  -  This nonane  solution contains  two
recovery standards, 13C12-1,2,3,4-TCDD and 13C12-l,2,3,7,8,9-HxCDD, at a nominal
concentration of 50 pg//iL  per compound.   10  to 50 juL of this solution will be
spiked into  each sample extract before the final concentration step and HRGC/HRMS
analysis.

      5.10  Matrix Spike Fortification Solution - Solution used to prepare the
MS and MSD samples.  It contains  all  unlabeled analytes listed in Table  5 at con-
centrations corresponding to the HRCC 3.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory material  to  this  chapter,  Organic  Analytes,
Sec. 4.1.

      6.2   Sample Collection

            6.2.1 Sample collection personnel  should,  to the extent possible,
      homogenize samples  in the field  before filling  the  sample  containers.
      This should minimize or eliminate the necessity for  sample homogenization
      in  the  laboratory.    The  analyst should  make  a  judgment, based  on the
      appearance of the sample, regarding the necessity for additional mixing.
      If the  sample  is  clearly  not  homogeneous,  the entire contents should be
      transferred to a glass or  stainless steel  pan  for  mixing with a stainless
      steel  spoon or spatula before removal of a sample portion for analysis.

            6.2.2 Grab  and  composite   samples must   be   collected   in  glass
      containers.  Conventional  sampling practices must be followed.  The bottle
      must not be prewashed with sample before collection.  Sampling equipment
      must be free of potential sources of contamination.

      6.3   Grinding or Blending of  Fish Samples - If not otherwise  specified by
the U.S. EPA, the whole fish (frozen) should be blended or ground to provide a
homogeneous sample.  The use of a stainless steel  meat  grinder with a 3 to 5 mm


                                   8290 -  10                         Revision 0
                                                                September 1994

-------
hole size inner plate  is recommended.  In some circumstances, analysis of fillet
or specific organs of fish may be requested by the U.S. EPA.  If so requested,
the above whole fish requirement is superseded.

      6.4   Storage and Holding  Times  - All  samples,  except  fish  and adipose
tissue samples, must be stored at 4°C in the dark, extracted within 30 days and
completely  analyzed  within 45 days  of extraction.    Fish and  adipose  tissue
samples  must  be  stored  at -20°C  in  the dark,  extracted within  30  days  and
completely analyzed within 45  days of collection.  Whenever samples  are analyzed
after the holding time expiration date, the results should be considered to be
minimum concentrations and should be identified as such.

      NOTE: The holding times  listed  in Sec. 6.4 are  recommendations.  PCDDs and
            PCDFs are very stable  in a  variety of matrices,  and holding times
            under the conditions  listed  in Sec. 6.4 may be as high as a year for
            certain  matrices.   Sample  extracts,  however,  should always  be
            analyzed within 45 days of extraction.

      6.5   Phase Separation - This is a guideline for phase separation for very
wet (>Z5 percent  water) soil, sediment  and paper pulp  samples.  Place  a 50 g
portion  in a  suitable centrifuge  bottle  and centrifuge  for 30 minutes  at
2,000 rpm.   Remove the  bottle  and  mark  the  interface  level  on  the  bottle.
Estimate the relative volume  of each phase.  With a disposable pipet,  transfer
the liquid  layer  into a clean bottle.   Mix the solid  with a  stainless steel
spatula  and remove  a portion to  be  weighed and analyzed  (percent dry  weight
determination, extraction).  Return the  remaining solid  portion to  the original
sample bottle (empty) or to a  clean sample bottle that  is properly  labeled, and
store  it as appropriate.   Analyze the  solid phase  by using only the soil,
sediment and paper pulp method.   Take note of,  and report, the estimated volume
of liquid before disposing of the liquid as a liquid waste.

      6.6   Soil,  Sediment,  or   Paper  Sludge   (Pulp)   Percent   Dry  Weight
Determination - The percent dry weight  of soil, sediment or paper pulp samples
showing detectable levels (see note below)  of at least one 2,3,7,8-substituted
PCDD/PCDF congener is determined  according  to  the following procedure.   Weigh a
10 g  portion  of the  soil  or  sediment  sample  (+  0.5 g)  to three  significant
figures.  Dry it to constant  weight at 110°C in an adequately  ventilated oven.
Allow  the  sample to  cool  in  a  desiccator.   Weigh  the dried solid  to three
significant figures.  Calculate and report the percent dry weight.   Do not use
this solid portion of the  sample  for extraction,  but  instead  dispose  of it as
hazardous waste.

      NOTE: Until detection limits have been established (Sec.  1.3),  the lower
            MCLs  (Table  1) may  be  used  to  estimate  the minimum detectable
            levels.

      % dry weight = q of dry sample x  100
                       g of sample

      CAUTION:    Finely  divided   soils   and  sediments   contaminated  with
                  PCDDs/PCDFs  are  hazardous   because   of  the  potential  for
                  inhalation  or ingestion of  particles containing  PCDDs/PCDFs
                                  8290  -  11                         Revision 0
                                                                September 1994

-------
                  (including 2,3,7,8-TCDD).   Such samples should be handled in
                  a confined environment (i.e.,  a closed hood or a glove box).

      6.7   Lipid Content Determination

            6.7.1 Fish Tissue - To determine the lipid content of fish tissue,
      concentrate 125 ml of the fish tissue extract (Sec.  7.2.2),  in a tared 200
      mL round bottom flask,  on a  rotary evaporator until a constant weight (W)
      is achieved.
                             100 (W)
            Percent lipid = 	
                               10

            Dispose of the lipid  residue as  a hazardous waste if the results of
      the analysis indicate the presence of PCDDs or PCDFs.

            6.7.2 Adipose Tissue  - Details for the determination of the adipose
      tissue lipid content are provided in Sec.  7.3.3.


7.0   PROCEDURE

      7.1   Internal standard addition

            7.1.1 Use a portion of 1 g to  1000 g  (± 5  percent) of the sample to
      be analyzed.  Typical  sample size requirements for different matrices are
      given in Sec. 7.4 and in Table 1.  Transfer the  sample portion to a tared
      flask and determine its weight.

            7.1.2 Except for adipose tissue, add  an appropriate quantity of the
      sample fortification mixture (Sec. 5.8) to the sample.  All samples should
      be spiked with 100 ^L of the sample fortification mixture to give internal
      standard concentrations as  indicated in Table 1.   As an example, for 13C12-
      2,3,7,8-TCDD, a 10 g soil sample requires the addition of 1000 pg of 13C12-
      2,3,7,8-TCDD to give the required 100 ppt fortification level.   The fish
      tissue sample (20 g) must be spiked  with 200 /iL of the internal  standard
      solution, because half of the extract  will  be used to determine the lipid
      content (Sec. 6.7.1).

                  7.1.2.1     For the fortification of soil, sediment, fly ash,
            water,  fish tissue,  paper  pulp and  wet  sludge  samples,  mix the
            sample fortification solution  with 1.0 ml acetone.

                  7.1.2.2     Do not dilute the  nonane  solution  for  the other
            matrices.

                  7.1.2.3     The fortification of adipose tissue is carried out
            at the time of homogenization (Sec.  7.3.2.3).

      7.2   Extraction  and Purification of Fish and Paper Pulp Samples

            7.2.1 Add  60  g anhydrous sodium  sulfate  to a  20 g portion  of a
      homogeneous  fish  sample  (Sec. 6.3)  and mix thoroughly  with  a  stainless


                                   8290  -  12                        Revision 0
                                                                September 1994

-------
steel spatula. After breaking up any lumps,  place the fish/sodium sulfate
mixture in the Soxhlet  apparatus on  top  of a glass wool plug.  Add 250 ml
methylene  chloride  or  hexane/methylene chloride  (1:1)  to  the  Soxhlet
apparatus  and  reflux  for 16 hours.   The solvent must  cycle completely
through the system five times per hour.  Follow the same procedure for the
partially  dewatered  paper pulp  sample   (using  a  10 g  sample,  30  g  of
anhydrous sodium sulfate and 200 ml of toluene).

      NOTE: As an  option,  a  Soxhlet/Dean Stark extractor  system  may be
            used, with  toluene as the solvent.   No sodium sulfate is added
            when using this option.

      7.2.2 Transfer  the fish  extract  from   Sec.  7.2.1  to  a  250  ml
volumetric flask and fill to  the mark with methylene chloride.  Mix well,
then  remove  125 ml  for the determination  of  the  lipid  content  (Sec.
6.7.1).   Transfer  the  remaining 125 ml of  the extract, plus  two  15 ml
hexane/methylene  chloride rinses   of   the  volumetric  flask,  to  a  KD
apparatus equipped with a Snyder column.  Quantitatively transfer all of
the paper pulp extract to a KD apparatus equipped with a Snyder column.

      NOTE: As an option, a rotary evaporator  may be used in place of the
            KD apparatus for the concentration of the extracts.

      7.2.3 Add a Teflon™, or equivalent, boiling chip.  Concentrate the
extract  in  a water bath to  an  apparent volume of  10  ml.   Remove the
apparatus from the water bath and allow to cool for 5 minutes.

      7.2.4 Add  50  mL  hexane and  a new boiling chip  to  the  KD  flask.
Concentrate in a water bath  to  an  apparent volume of 5 mL.   Remove the
apparatus from the water bath and allow to cool for 5 minutes.

      NOTE: The  methylene  chloride  must have  been  completely  removed
            before proceeding with  the next step.

      7.2.5 Remove and invert the Snyder column and rinse it into the KD
apparatus with two 1 ml portions of hexane.   Decant the contents of the KD
apparatus and concentrator tube  into a  125  mL separatory  funnel.   Rinse
the KD apparatus with two additional 5 mL portions  of hexane and add the
rinses  to the  funnel.    Proceed   with the  cleanup according  to  the
instructions starting  in Sec. 7.5.1.1,  but omit the procedures described
in Sees. 7.5.1.2 and 7.5.1.3.

7.3   Extraction and Purification of Human Adipose  Tissue

      7.3.1 Human adipose tissue samples must  be stored at a temperature
of -20°C or lower  from  the time of collection until  the time of analysis.
The use of chlorinated  materials during the  collection of the samples must
be avoided.  Samples are handled with stainless steel  forceps, spatulas,
or scissors.   All sample bottles (glass) are cleaned  as  specified in the
note at the end of Sec. 4.3.   Teflon™ lined caps  should be used.
                            8290  -  13                         Revision 0
                                                          September 1994

-------
NOTE: The specified  storage temperature  of -20°C  is  the  maximum
      storage temperature permissible  for adipose  tissue samples.
      Lower storage temperatures are recommended.

7.3.2 Adipose Tissue Extraction

      7.3.2.1     Weigh, to the nearest 0.01 g,  a 10 g portion of a
frozen adipose tissue sample into a culture tube (2.2 x 15 cm).

      NOTE:  The   sample   size  may   be  smaller,    depending   on
            availability.  In  such  a  situation,   the  analyst  is
            required to adjust the volume of the internal standard
            solution added to the sample to meet the fortification
            level stipulated in Table 1.

      7.3.2.2     Allow the  adipose tissue  specimen  to reach room
temperature (up to 2 hours).

      7.3.2.3     Add 10 ml  methylene  chloride and 100  /*L  of the
sample  fortification   solution.     Homogenize  the  mixture  for
approximately 1 minute with a tissue homogenizer.

      7.3.2.4     Allow the  mixture  to separate,  then  remove the
methylene chloride extract from the residual solid material  with a
disposable pi pet. Percolate the methylene  chloride  through a filter
funnel containing a clean glass wool  plug  and 10 g  anhydrous sodium
sulfate.  Collect the dried extract in a graduated 100 ml volumetric
flask.

      7.3.2.5     Add a second 10 ml portion of methylene chloride
to the sample and homogenize for 1 minute.   Decant  the solvent, dry
it, and transfer  it to the 100 ml volumetric flask (Sec. 7.3.2.4).

      7.3.2.6     Rinse  the   culture   tube with   at   least  two
additional portions of methylene chloride (10 ml each), and transfer
the entire contents to  the  filter  funnel  containing the anhydrous
sodium sulfate.   Rinse  the  filter  funnel  and  the anhydrous sodium
sulfate contents  with additional methylene  chloride  (20 to 40 ml)
into the 100 ml flask.   Discard the sodium  sulfate.

      7.3.2.7     Adjust  the  volume  to   the   100 ml  mark  with
methylerie chloride.

7.3.3 Adipose Tissue Lipid Content Determination

      7.3.3.1     Preweigh  a clean  1 dram  (or metric  equivalent)
glass vial to the nearest 0.0001 g on an analytical  balance tared to
zero.

      7.3.3.2     Accurately transfer  1.0 ml of the final extract
(100 mL) from Sec.  7.3.2.7  to  the  vial.   Reduce the volume of the
extract on  a water bath (50-60°C)  by  a gentle stream  of purified
                       8290 - 14                        Revision 0
                                                    September 1994

-------
nitrogen  until  an  oily  residue  remains.   Nitrogen  blowdown  is
continued until a constant weight is achieved.

      NOTE: When the  sample  size  of the adipose tissue is  smaller
            than 10 g, then the analyst may use a larger portion (up
            to  10 percent) of  the extract defined in Sec.  7.3.2.7
            for the lipid determination.

      7.3.3.3     Accurately weigh the vial  with the  residue to the
nearest 0.0001  g and  calculate  the  weight of the lipid present  in
the vial based  on the difference  of the weights.

      7.3.3.4     Calculate  the  percent  lipid  content   of   the
original sample to the nearest 0.1 percent as shown  below:

                               **lr  X  "ext
      Lipid content,  LC {%)  = 	     x   100
                               """at X "al
where:

      W,r    =     weight of the lipid residue to the  nearest 0.0001
                  g calculated from Sec. 7.3.3.3,

      Vext   =     total volume  (100  ml)  of  the extract in  ml  from
                  Sec. 7.3.2.7,

      Wat    =     weight  of  the  original  adipose tissue sample  to
                  the nearest 0.01 g from Sec.  7.3.2.1, and

      Val    =     volume of the aliquot of the  final extract in  mi
                  used  for  the quantitative measure  of the lipid
                  residue (1.0 ml) from Sec. 7.3.3.2.

      7.3.3.5     Record the lipid residue measured  in  Sec.  7.3.3.3
and the percent lipid content from Sec. 7.3.3.4.

7.3.4 Adipose Tissue  Extract Concentration

      7.3.4.1     Quantitatively transfer the remaining extract  from
Sec.  7.3.3.2  (99.0  ml)  to a 500  ml Erlenmeyer  flask.   Rinse the
volumetric flask with 20 to 30 ml of additional methylene chloride
to ensure quantitative transfer.

      7.3.4.2     Concentrate the extract on a rotary evaporator and
a water bath at 40°C until an oily residue remains.

7.3.5 Adipose Tissue  Extract Cleanup

      7.3.5.1     Add 200 ml hexane  to  the lipid residue in  the 500
ml Erlenmeyer flask and swirl the flask to dissolve the residue.
                       8290  -  15                         Revision 0
                                                    September 1994

-------
            7.3.5.2     Slowly add, with  stirring,  100 g of  40  percent
      (w/w) sulfuric acid-impregnated silica gel.   Stir  with  a magnetic
      stirrer for two hours at room temperature.

            7.3.5.3     Allow the solid phase to  settle,  and  decant the
      liquid through  a  filter funnel containing  10 g anhydrous  sodium
      sulfate on a glass  wool  plug,  into another 500 mL Erlenmeyer flask.

            7.3.5.4     Rinse the solid  phase with two 50 ml portions of
      hexane.   Stir  each rinse  for 15  minutes,  decant,  and  dry  as
      described under Sec.  7.3.5.3.  Combine the hexane extracts from Sec.
      7.3.5.3 with the rinses.

            7.3.5.5     Rinse the sodium sulfate in the filter  funnel with
      an additional 25 ml  hexane  and combine this  rinse  with  the hexane
      extracts from Sec.  7.3.5.4.

            7.3.5.6     Prepare an acidic  silica column as follows:  Pack
      a 2 cm x 10  cm  chromatographic column with  a glass wool plug, add
      approximately 20 ml hexane,  add 1 g  silica gel and allow to settle,
      then add 4 g of 40  percent  (w/w)  sulfuric  acid-impregnated silica
      gel and allow to settle.   Elute the excess  hexane  from  the column
      until  the  solvent   level  reaches  the  top  of the  chromatographic
      packing.  Verify that the column does  not  have any  air bubbles and
      channels.

            7.3.5.7     Quantitatively  transfer the hexane  extract from
      the Erlenmeyer flask (Sees. 7.3.5.3  through 7.3.5.5) to  the silica
      gel column reservoir. Allow the hexane extract to percolate through
      the column and collect the eluate  in a 500 mL KD apparatus.

            7.3.5.8     Complete the elution by  percolating 50 ml hexane
      through the column  into  the KD apparatus.  Concentrate the eluate on
      a steam bath to approximately 5 ml.  Use nitrogen blowdown to bring
      the final volume to about 100 juL.

            NOTE:  If the silica gel impregnated  with 40 percent sulfuric
                   acid is highly discolored throughout the length of the
                   adsorbent bed,  the cleaning procedure must be repeated
                   beginning with Sec.  7.3.5.1.

            7.3.5.9     The  extract is  ready  for  the  column  cleanups
      described in Sees.  7.5.2 through  7.5.3.6.

7.4   Extraction and Purification of Environmental and Waste Samples

      7.4.1 Sludge/Wet Fuel Oil

            7.4.1.1     Extract aqueous  sludge or  wet fuel oil samples by
      refluxing a sample (e.g., 2 g) with  50 ml toluene in a 125 ml flask
      fitted with  a  Dean-Stark water separator.   Continue refluxing the
      sample until all the water  is removed.
                             8290  -  16                         Revision 0
                                                          September 1994

-------
      NOTE: If the sludge or fuel oil sample dissolves in toluene,
            treat  it  according  to the  instructions  in Sec. 7.4.2
            below.   If the labeled  sludge  sample  originates from
            pulp   (paper   mills),   treat   it   according   to  the
            instructions  starting in  Sec. 7.2,  but  without  the
            addition of sodium sulfate.

      7.4.1.2     Cool  the  sample,   filter the  toluene  extract
through a  glass  fiber filter,  or equivalent,  into  a 100  mL round
bottom flask.

      7.4.1.3     Rinse the filter with  10 ml  toluene and combine
the extract with the rinse.

      7.4.1.4     Concentrate the combined solutions to near dryness
on a rotary evaporator at 50°C.   Use  of an  inert gas to concentrate
the extract is also permitted.  Proceed with Sec. 7.4.4.

7.4.2 Still Bottom/Oil

      7.4.2.1     Extract still  bottom or  oil  samples by  mixing a
sample portion (e.g.,  1.0 g) with  10 ml  toluene in a small  beaker
and  filtering the  solution  through  a  glass  fiber filter  (or
equivalent) into a 50 ml round bottom flask.  Rinse the beaker and
filter with 10 ml toluene.

      7.4.2.2     Concentrate the combined toluene solutions  to near
dryness on a rotary evaporator at 50°C.   Proceed with Sec.  7.4.4.
7.4.3 Fly Ash

NOTE: Because of the  tendency  of fly ash  to "fly",   all  handling
      steps  should  be performed  in  a hood in order  to  minimize
      contamination.

      7.4.3.1     Weigh about  10  g fly ash to two decimal places and
transfer to  an  extraction jar.   Add 100 /xL  sample fortification
solution (Sec. 5.8), diluted to  1 ml with  acetone,  to the sample.
Add 150 ml of 1 M HC1  to  the fly  ash  sample.  Seal the jar with the
Teflon™ lined screw cap  and shake for 3 hours at room temperature.

      7.4.3.2     Rinse  a  glass  fiber filter  with  toluene,  and
filter the  sample  through the filter  paper,  placed  in  a Buchner
funnel, into all  flask.  Wash the fly ash  cake with approximately
500 ml organic-free reagent water and dry the filter cake overnight
at room temperature in a desiccator.

      7.4.3.3     Add  10 g anhydrous powdered  sodium sulfate,  mix
thoroughly, let sit in a closed container for one hour, mix again,
let sit for another hour, and mix again.

      7.4.3.4     Place the  sample  and the filter paper  into  an
extraction thimble, and extract  in a Soxhlet  extraction apparatus
                       8290  -  17                         Revision 0
                                                    September 1994

-------
      charged with  200 ml  toluene for 16 hours using  a  five cycle/hour
      schedule.

            NOTE: As an option,  a Soxhlet/Dean Stark extractor system may
                  be used,  with  toluene as the solvent.  No sodium sulfate
                  is added when using this option.

            7.4.3.5     Cool  and  filter the  toluene  extract  through a
      glass fiber  filter  into a  500  ml  round bottom flask.   Rinse the
      filter  with   10 ml  toluene.   Add  the  rinse to  the  extract and
      concentrate  the  combined  toluene  solutions  to  near  dryness  on a
      rotary evaporator at 50°C.   Proceed with Sec.  7.4.4.

      7.4.4 Transfer the concentrate to a 125 ml separatory funnel  using
15 ml hexane.  Rinse the flask with  two  5  ml  portions  of hexane and add
the rinses to the funnel.   Shake  the  combined solutions in the separatory
funnel for two minutes with 50 ml of 5 percent sodium chloride solution,
discard the aqueous layer, and proceed with Sec. 7.5.

      7.4.5 Aqueous samples

            7.4.5.1     Allow the sample to come to ambient temperature,
      then mark  the water  meniscus on  the  side  of the  1  L sample bottle
      for  later  determination  of the  exact  sample  volume.    Add the
      required acetone diluted sample fortification solution (Sec.  5.8).

            7.4.5.2     When the sample is judged  to contain  1 percent or
      more  solids,  the sample  must  be  filtered  through a  glass  fiber
      filter that has been rinsed with toluene.  If the suspended solids
      content  is  too  great  to filter  through  the  0.45 /urn  filter,
      centrifuge the sample, decant,  and  then  filter the  aqueous  phase.

            NOTE: Paper mill effluent samples  normally contain 0.02%-0.2%
                  solids, and would  not require  filtration.  However, for
                  optimum  analytical  results,  all  paper  mill  effluent
                  samples  should be  filtered,  the isolated  solids and
                  filtrate  extracted   separately,  and   the  extracts
                  recombined.

            7.4.5.3     Combine the solids from the centrifuge bottle(s)
      with the particulates on the filter and with the filter itself and
      proceed with  the  Soxhlet  extraction  as  specified in Sees. 7.4.6.1
      through 7.4.6.4.  Remove and invert the Snyder column and rinse it
      down into  the KD apparatus  with two 1 ml portions of hexane.

            7.4.5.4     Pour  the  aqueous  filtrate  into a 2 L separatory
      funnel.  Add 60  ml methylene chloride to the sample  bottle,  seal and
      shake  for  30 seconds to  rinse the  inner surface.   Transfer the
      solvent to the separatory funnel and extract the sample by shaking
      the funnel for two minutes  with periodic venting.

            7..4.5.S     Allow the organic layer to separate from the  water
      phase  for  a  minimum of  10 minutes.    If the emulsion  interface


                             8290 -  18                        Revision 0
                                                          September 1994

-------
between  layers  is more than one  third the volume  of the solvent
layer, the analyst must employ mechanical techniques  to complete the
phase separation  (e.g., glass stirring rod).

      7.4.5.6     Collect the methylene chloride into a KD apparatus
(mounted  with  a  10  ml concentrator  tube)  by passing  the sample
extracts through  a filter funnel packed with a glass wool plug and
5 g anhydrous sodium sulfate.

      NOTE: As an option, a rotary evaporator may be used in place
            of  the  KD apparatus  for  the concentration  of  the
            extracts.

      7.4.5.7     Repeat  the extraction  twice with  fresh  60  ml
portions of methylene chloride.  After the third extraction, rinse
the sodium sulfate with  an  additional  30 ml methylene chloride to
ensure quantitative transfer.  Combine all extracts and the rinse in
the KD apparatus.

      NOTE: A continuous  liquid-liquid  extractor  may be  used  in
            place of  a separatory  funnel  when experience  with  a
            sample from  a given  source  indicates  that  a  serious
            emulsion  problem  will   result   or   an  emulsion  is
            encountered when using a separatory funnel.   Add 60 ml
            methylene chloride  to the sample bottle,  seal, and shake
            for 30 seconds to rinse the inner surface.  Transfer the
            solvent  to  the  extractor.   Repeat  the rinse  of  the
            sample bottle with  an additional 50 to 100  ml portion of
            methylene chloride and add the rinse to the extractor.
            Add 200 to 500 ml methylene chloride to the distilling
            flask, add sufficient organic-free reagent water (Sec.
            5.1)  to  ensure proper  operation,  and  extract  for
            24 hours.  Allow to cool, then detach  the  distilling
            flask.  Dry and  concentrate the extract  as described in
            Sees. 7.4.5.6 and  7.4.5.8 through 7.4.5.10.   Proceed
            with  Sec. 7.4.5.11.

      7.4.5.8     Attach a Snyder column and concentrate the extract
on a  water bath  until  the apparent volume of  the  liquid is 5 ml.
Remove the KD apparatus and  allow  it to drain and cool for at least
10 minutes.

      7.4.5.9     Remove the Snyder column, add  50  mL hexane,  add
the  concentrate  obtained  from the   Soxhlet  extraction  of  the
suspended solids  (Sec.  7.4.5.3), if applicable,  re-attach the Snyder
column,  and concentrate to  approximately  5 ml.   Add a new boiling
chip  to  the KD  apparatus before  proceeding  with  the  second
concentration step.

      7.4.5.10    Rinse the flask  and  the lower joint with two 5 mL
portions of hexane and combine  the rinses with the extract to give
a final  volume of about 15 ml.
                       8290  -  19                         Revision 0
                                                    September 1994

-------
            7.4.5.11    Determine the original  sample  volume  by filling
      the sample bottle to  the mark with water and transferring the water
      to a 1000 ml graduated cylinder.  Record  the  sample  volume to the
      nearest 5 ml.  Proceed with Sec. 7.5.

      7.4.6 Soil/Sediment

            7.4.6.1     Add 10  g anhydrous powdered sodium sulfate to the
      sample portion  (e.g.,  10 g) and mix  thoroughly with a stainless
      steel spatula.  After breaking up any lumps,  place the soil/sodium
      sulfate mixture in the Soxhlet apparatus on top of a glass wool plug
      (the use of an extraction thimble is optional).

            NOTE: As an option,  a Soxhlet/Dean Stark extractor system may
                  be used,  with  toluene as the solvent.  No sodium sulfate
                  is added when using this option.

            7.4.6.2     Add 200  to 250 ml toluene to  the Soxhlet apparatus
      and reflux for 16 hours.   The solvent must cycle completely through
      the system five times per hour.

            NOTE: If the dried  sample is not of  free flowing consistency,
                  more sodium sulfate must be added.

            7.4.6.3     Cool and filter the extract through a glass fiber
      filter into  a 500 ml  round bottom flask for evaporation  of the
      toluene.   Rinse  the  filter  with  10  ml  of  toluene,  and concentrate
      the combined  fractions to near dryness on a  rotary  evaporator at
      50°C.   Remove the flask from the water  bath and  allow to cool for
      5 minutes.

            7.4.6.4     Transfer  the  residue   to  a  125  mL  separatory
      funnel, using 15 mL of hexane.   Rinse the  flask with two additional
      portions of hexane, and add the rinses to  the funnel.  Proceed with
      Sec. 7.5.

7.5   Cleanup

      7.5.1 Partition

            7.5.1.1     Partition the hexane extract  against  40 ml of
      concentrated  sulfuric  acid.   Shake  for two minutes.   Remove and
      discard the sulfuric acid layer (bottom).   Repeat the acid washing
      until no color  is  visible in  the acid  layer  (perform a  maximum of
      four acid washings).

            7.5.1.2     Omit this  step  for  the fish sample  extract.
      Partition  the  extract against  40  ml  of  5 percent  (w/v)  aqueous
      sodium chloride.   Shake  for two minutes.   Remove  and discard the
      aqueous layer (bottom).

            7,.5.1.3     Omit this  step  for  the fish sample  extract.
      Partition  the  extract against  40 ml  of  20 percent  (w/v)  aqueous


                             8290 -  20                        Revision 0
                                                          September  1994

-------
potassium  hydroxide (KOH).   Shake for  two minutes.    Remove and
discard the aqueous layer  (bottom).  Repeat the base washing until
no color is visible in the bottom layer (perform a maximum of four
base  washings).   Strong  base (KOH)  is  known to  degrade certain
PCDDs/PCDFs, so contact time must be minimized.

      7.5.1.4     Partition the extract against 40 ml of 5 percent
(w/v) aqueous sodium chloride.  Shake for two minutes.  Remove and
discard the aqueous layer  (bottom).  Dry the extract by pouring it
through a  filter  funnel  containing anhydrous  sodium  sulfate  on a
glass wool  plug,  and  collect it in  a  50 ml  round  bottom flask.
Rinse the funnel  with the sodium  sulfate with  two 15 ml portions of
hexane, add  the  rinses  to the 50  ml flask,  and  concentrate the
hexane solution to  near dryness on  a  rotary evaporator  (35°C water
bath), making  sure  all  traces of  toluene (when  applicable)  are
removed.    (Use  of  blowdown with  an inert gas to  concentrate the
extract is also permitted.)

7.5.2 Silica/Alumina Column Cleanup

      7.5.2.1     Pack a gravity  column  (glass,  30 cm x 10.5 mm),
fitted with a Teflon™  stopcock, with silica gel as  follows:  Insert
a glass wool plug into the bottom of the column.   Place 1 g silica
gel  in the  column  and tap the column gently  to  settle  the silica
gel.  Add 2 g sodium hydroxide-impregnated  silica gel, 4 g sulfuric
acid-impregnated silica  gel,  and  2 g  silica gel.   Tap  the column
gently after each addition.   A small  positive pressure  (5 psi)  of
clean nitrogen may  be used if needed.   Elute with 10 ml hexane and
close the stopcock  just before exposure of the top layer of silica
gel to air.  Discard the eluate.   Check the column for channeling.
If  channeling  is  observed, discard the  column.    Do not  tap the
wetted column.

      7.5.2.2     Pack a gravity  column (glass, 300 mm x 10.5 mm),
fitted with a Teflon™ stopcock, with  alumina  as follows:  Insert a
glass wool plug into the bottom of the column.  Add a 4 g layer of
sodium sulfate.  Add a 4  g  layer of Woelm® Super 1 neutral alumina.
Tap the top of the  column  gently.   Woelm®  Super  1  neutral alumina
need not  be activated or cleaned before use,  but it  should be stored
in a sealed desiccator.  Add a 4 g  layer of anhydrous sodium sulfate
to  cover  the  alumina.    Elute with  10  mL hexane  and   close  the
stopcock just before exposure  of  the  sodium sulfate layer to air.
Discard the eluate.  Check the column for channeling.  If channeling
is observed, discard the column.   Do not  tap a wetted column.

      NOTE: Optionally, acidic alumina (Sec. 5.2.2) can  be used in
            place of neutral  alumina.

      7.5.2.3     Dissolve the residue  from Sec.  7.5.1.4  in  2 ml
hexane and apply the hexane solution  to the  top  of the  silica gel
column.  Rinse the flask with enough hexane  (3-4 ml) to complete the
quantitative transfer  of the  sample to the surface of  the silica
gel.


                       8290 -  21                         Revision 0
                                                    September 1994

-------
      7.5.2.4     Elute the silica  gel column with 90 ml of hexane,
concentrate the eluate on a rotary evaporator (35°C  water bath) to
approximately 1 ml,  and apply the  concentrate  to the top  of the
alumina column  (Sec. 7.5.2.2).   Rinse the  rotary evaporator flask
twice with  2  ml of hexane, and  add  the  rinses  to the top  of the
alumina column.

      7.5.2.5     Add 20 ml hexane to the alumina column  and elute
until the hexane level  is  just below  the top of the sodium sulfate.
Do not discard the eluted hexane, but collect it  in a separate flask
and store it for later use, as it may be useful in determining where
the  labeled  analytes  are  being  lost  if  recoveries   are  not
satisfactory.

      7.5.2.6     Add  15  ml  of  60  percent methylene  chloride in
hexane  (v/v)  to the alumina  column and collect  the eluate  in  a
conical  shaped   (15  ml)  concentration  tube.    With  a  carefully
regulated stream of nitrogen,  concentrate the 60 percent  methylene
chloride/hexane fraction to about 2 ml.

7.5.3 Carbon Column Cleanup

      7.5.3.1     Prepare an AX-21/Celite  545®  column  as follows:
Thoroughly mix 5.40 g active carbon AX-21 and 62.0 g Celite 545® to
produce an 8 percent (w/w) mixture.  Activate the mixture at  130°C
for 6 hours and store it in a desiccator.

      7.5.3.2     Cut  off  both  ends   of  a   10  ml   disposable
serological pipet to give  a  10 cm long column.   Fire  polish both
ends and flare, if desired.   Insert  a glass wool  plug  at one end,
then pack the column with enough Celite  545® to form a 1 cm plug,
add 1 g of the AX-21/Celite 545® mixture,  top with additional Celite
545®  (enough  for  a  1  cm  plug),  and cap the packing with another
glass wool plug.

      NOTE: Each new batch of AX-21/Celite 545® must be checked as
            follows:    Add  50 juL  of the   continuing  calibration
            solution to 950 juL hexane.  Take this solution through
            the carbon  column  cleanup step, concentrate  to 50 inl-
            and analyze.  If the recovery of any  of the analytes is
            <80 percent, discard this batch of AX-21/Celite 545®.

      7.5.3.3     Rinse the AX-21/Celite 545® column with 5 ml of
toluene,  followed  by  2 ml  of  75:20:5  (v/v) methylene  chloride/
methanol/toluene, 1 mL of  1:1  (v/v) cyclohexane/methylene chloride,
and  5 ml  hexane.   The  flow  rate should be  less  than  0.5 mL/min.
Discard the rinses.  While the column is  still wet with hexane, add
the  sample  concentrate (Sec.  7.5.2.6)  to  the  top  of  the column.
Rinse the concentrator  tube (which contained the  sample concentrate)
twice with 1 ml  hexane,  and add the rinses to the top  of the column.

      7.5.3.4     Elute  the  column  sequentially  with  two  2  mL
portions  of hexane,  2 ml cyclohexane/methylene  chloride  (50:50,


                       8290 -  22                        Revision 0
                                                    September  1994

-------
            v/v), and 2 ml methylene chloride/methanol/toluene (75:20:5, v/v).
            Combine these eluates;  this combined fraction may be used  as  a check
            on column efficiency.

                  7.5.3.5     Turn  the  column  upside  down  and  elute  the
            PCDD/PCDF fraction with 20  ml toluene.  Verify that no carbon fines
            are  present  in the  eluate.   If carbon  fines  are present  in the
            eluate, filter the eluate through a glass  fiber filter  (0.45 urn) and
            rinse the filter with 2 ml toluene.  Add the rinse to the eluate.

                  7.5.3.6     Concentrate the toluene fraction  to  about  1 ml on
            a  rotary  evaporator  by  using  a  water  bath  at 50°C.   Carefully
            transfer the concentrate into a 1 ml minivial  and, again at elevated
            temperature (50°C), reduce the volume to about 100 pi using a stream
            of nitrogen  and  a sand bath.   Rinse the  rotary  evaporator flask
            three  times  with  300  p.1  of  a  solution  of  1  percent toluene  in
            methylene chloride,  and  add the  rinses  to the concentrate.   Add
            10 juL of the nonane recovery standard solution (Sec. 5.9) for soil,
            sediment,  water,  fish,  paper pulp  and adipose tissue samples, or 50
            juL of  the recovery standard  solution for sludge,  still  bottom and
            fly ash samples.   Store the  sample  at room  temperature in the dark.

      7.6   Chromatographic/Mass Spectrometric Conditions and Data Acquisition
Parameters

            7.6.1 Gas Chromatograph

            Column coating:         DB-5
            Film thickness:         0.25 urn
            Column dimension:       60 m x 0.32 mm
            Injector temperature:   270°C
            Splitless valve time:   45 s
            Interface temperature:  Function of the final temperature
            Temperature program:
Stage


Init.
Temp.
(°C)
Init.
Hold Time
(min)
Temp.
Ramp
(°C/min)
Final
Temp.
(°C)
Final
Hold
Time (min)
             1        200       2           5            220          16
             2                             5            235          7
             3                             5            330          5

            Total  time:  60 min

            7.6.2  Mass  Spectrometer

                  7.6.2.1     The  mass  spectrometer  must  be  operated  in  a
            selected  ion   monitoring   (SIM)  mode  with  a  total  cycle  time
            (including  the  voltage reset  time)  of one  second  or  less  (Sec.
            7.6.3.1).   At  a minimum, the ions listed in Table 6 for each of the
            five  SIM  descriptors  must be monitored.    Note  that  with  the


                                   8290 - 23                         Revision 0
                                                                September 1994

-------
      exception  of  the  last descriptor  (OCDD/OCDF),  all   descriptors
      contain  10 ions.   The selection (Table 6) of the molecular  ions M
      and  M+2  for 13C-HxCDF and  13C-HpCDF rather  than  M+2  and M+4  (for
      consistency) was made to eliminate, even under high-resolution  mass
      spectrometric  conditions,  interferences  occurring  in these two ion
      channels for samples containing high  levels of native HxCDDs  and
      HpCDDs.   It  is important to maintain the  same  set  of ions for  both
      calibration  and sample extract analyses.  The selection of the lock-
      mass ion is  left  to the performing laboratory.

            NOTE:       At the  option  of the  analyst,  the  tetra-  and
                        pentachlorinated  dioxins   and  furans  can  be
                        combined into a  single  descriptor.

            7.6.2.2     The   recommended    mass    spectrometer    tuning
      conditions are based on the groups of monitored ions shown  in Table
      6.  By using  a PFK molecular leak, tune the  instrument  to meet the
      minimum  required  resolving power of 10,000  (10 percent valley) at
      m/z  304.9824  (PFK) or any other  reference signal close  to  m/z
      303.9016 (from TCDF).  By using  peak  matching conditions and  the
      aforementioned PFK reference peak, verify that the exact mass of m/z
      380.9760 (PFK) is  within 5 ppm of the required value.  Note that the
      selection  of  the  low-  and high-mass  ions must  be such that  they
      provide  the  largest voltage jump  performed  in  any  of the  five  mass
      descriptors  (Table 6).

      7.6.3 Data Acquisition

            7.6.3.1     The total  cycle  time for data acquisition must be
      < 1  second.  The total cycle time  includes the sum of all  the dwell
      times and  voltage  reset times.

            7.6.3.2     Acquire  SIM data for all  the ions listed  in the
      five descriptors  of Table  6.

7.7   Calibration

      7.7.1 Initial Calibration  - Initial calibration is required  before
any samples are  analyzed for PCDDs and PCDFs.  Initial calibration is  also
required  if  any  routine calibration  (Sec. 7.7.3)  does not  meet  the
required criteria listed in Sec. 7.7.2.

            7.7.1.1     All five high-resolution concentration calibration
      solutions   listed  in  Table  5  must   be  used  for   the   initial
      calibration.

            7.7.1.2    Tune  the  instrument with  PFK  as  described  in
      Sec. 7.6.2.2.

            7.7.1.3    Inject  2 fil  of  the  GC column performance check
      solution (Sec. 5.7) and acquire SIM mass spectral data as described
                            8290  - 24                         Revision 0
                                                          September 1994

-------
earlier in Sec.  7.6.2.  The total cycle time must be <  1 second.  The
laboratory must not perform  any  further analysis until it is demon-
strated and documented that  the criterion listed in Sec. 8.2.1 was
met.

       7.7.1.4     By  using   the  same  GC  (Sec.   7.6.1)  and  MS
(Sec. 7.6.2) conditions that produced acceptable  results with the
column performance check solution, analyze a  2  /LtL portion of each
of  the five  concentration  calibration  solutions  once with  the
following mass spectrometer  operating parameters.

            7.7.1.4.1    The  ratio of  integrated  ion current for the
       ions  appearing  in Table  8 (homologous  series quantitation
       ions)  must be within  the  indicated  control  limits (set for
       each   homologous  series)   for  all   unlabeled  calibration
       standards  in Table 5.

            7.7.1.4.2    The  ratio of  integrated  ion current for the
       ions  belonging  to the carbon-labeled  internal  and recovery
       standards   (Table  5)   must  be within  the  control  limits
       stipulated in Table 8.

            NOTE: Sees. 7.7.1.4.1 and 7.7.1.4.2  require  that 17 ion
                  ratios from  Sec. 7.7.1.4.1 and 11 ion ratios from
                  Sec. 7.7.1.4.2  be  within the specified  control
                  limits simultaneously  in one  run.   It  is  the
                  laboratory's  responsibility to   take  corrective
                  action if the  ion  abundance  ratios  are  outside
                  the limits.

            7.7.1.4.3    For  each  selected ion current  profile (SICP)
       and for each GC signal  corresponding  to the elution  of  a
       target  analyte  and of its  labeled  standards, the signal-to-
       noise  ratio  (S/N)  must  be better  than or  equal to  2.5.
       Measurement of  S/N  is required for any GC  peak  that  has an
       apparent S/N of less than 5:1.   The result of the calculation
       must  appear on  the SICP  above  the  GC peak in question.

            7.7.1.4.4    Referring to Table  9,  calculate  the  17
       relative response factors  (RF) for unlabeled target analytes
       [RF(n);  n  = 1  to 17]  relative  to  their appropriate internal
       standards  (Table 5)  and  the  nine  RFs  for  the  labeled  13C12
       internal standards [RF(m); m =  18 to  26)]  relative to the two
       recovery   standards  (Table 5)  according to  the  following
       formulae:
                      8290  - 25                         Revision 0
                                                    September 1994

-------
             Ax x Qjs                       Ais x Qre
  RFn    =  	            RFm    =  	
             Qx * Ais                       Qi8 x Are

where:

     Ax     =     sum of the integrated ion abundances of the
                  quantitation  ions  (Tables 6  and  9)  for
                  unlabeled  PCDDs/PCDFs,

     Ais     =     sum of the integrated ion abundances of the
                  quantitation  ions  (Tables  6 and  9)  for the
                  labeled  internal  standards,

     Ars     =     sum of the integrated ion abundances of the
                  quantitation  ions  (Tables  6 and  9)  for the
                  labeled  recovery  standards,

     Qis     =     quantity of the  internal  standard injected
                  (pg),

     Qr8     =     quantity of the  recovery  standard injected
                  (pg),  and

     Qx     =     quantity of the unlabeled PCDD/PCDF analyte
                  injected (pg).

     The RFn and RFm are dimensionless quantities; the units
used to express Qis, Qre and Qx must  be the same.
     7.7.1.4.5    Calculate  the  RF  and  their  respective
percent  relative standard  deviations  (%RSO)  for the  five
calibration solutions:
                    5
     RFn   - 1/5  Z RFnW
     Where  n  represents  a  particular PCDD/PCDF  (2,3,7,8-
substituted) congener  (n  =  1 to 17; Table 9),  and  j  is the
injection number  (or  calibration solution number;  j  =  1 to
5).

     7.7.1.4.6    The  relative response factors to be used for
the determination of the concentration of total isomers in a
homologous series (Table 9)  are  calculated as  follows:
                8290 - 26                         Revision 0
                                              September 1994

-------
      7.7.1.4.6.1  For   congeners  that  belong  to  a
homologous series containing only one isomer (e.g., OCDD
and  OCDF)  or  only  one  2,3,7,8-substituted   isomer
(Table 4; TCDD,  PeCDD,  HpCDD,  and_JCDF),  the  mean  RF
used will be the same as the mean RF determined  in Sec.
7.7.1.4.5.

      NOTE: The  calibration  solutions  do  not  contain
            13C12-OCDF as an internal  standard.   This is
            because a minimum resolving power of 12,000
            is  required to resolve the  [M+6]+  ion of
            13C12-OCDF from the [M+2]+  ion  of OCDD (and
            [M+4]+  from 13C12lOCDF  with [M]+ of OCDD).
            Therefore,  the  RF for  OCDF is  calculated
            relative to  13C12-OCDD.

      7.7.1.4.6.2  For   congeners  that  belong  to  a
homologous    series    containing   more    than_  one
2,3,7,8-substituted isomer (Table 4), the mean  RF used
for those homologous series will  be the  mean  of  the RFs
calculated  for   all   individual  2,3,7,8-substituted
congeners using the equation  below:

                   1       t
      RFU   =      -      I RFn
        k

                   t      n=1
where:
       k  =  27  to  30  (Table 9),  with 27 = PeCDF; 28 =
            HxCDF; 29 = HxCDD; and 30 = HpCDF,

       t  =  total number of 2,3,7,8-substituted isomers
            present in the calibration solutions (Table
            5)  for each homologous  series (e.g., two
            for PeCDF, four for HxCDF, three for HxCDD,
            two for HpCDF).

       NOTE: Presumably, the HRGC/HRMS response  factors
            of  different  isomers within  a homologous
            series  are   different.     However,  this
            analytical    protocol    will    make   the
            assumption that the HRGC/HRMS  responses of
            all isomers in  a homologous series  that do
            not have  the  2,3,7,8-substitution  pattern
            are the  same as  the responses of one or
            more of  the 2,3,7,8-substituted  isomer(s)
            in that homologous series.
           8290 - 27                        Revision 0
                                        September  1994

-------
                  7.7.1.4.7    Relative response factors [RFm] to be used
            for the determination of the percent recoveries for the nine
            internal  standards  are  calculated  as  follows:
                  RF
     Qism  *  Are

               5
RFm  = 1/5    I  RFm(j)
                                            m(j),
            where:

                  m      =     18 to  26  (congener type)  and  j =  1  to 5
                              (injection number),

                  Aism    =     sum of the integrated  ion abundances of the
                              quantitation ions  (Tables  6 and 9)  for a
                              given internal  standard (m - 18 to 26),

                  Ars    =     sum of the integrated  ion abundances of the
                              quantitation ions (Tables 6 and 9)  for the
                              appropriate recovery standard  (see Table 5,
                              footnotes),

                  Qrs>  QiSm =    quantities of, respectively,  the recovery
                              standard  (rs)  and  a  particular  internal
                              standard (is = m) injected (pg),

                  RFm  =        relative  response  factor of  a particular
                              internal   standard   (m)   relative   to  an
                              appropriate    recovery    standard,    as
                              determined from one injection, and

                  RFm  =        calculated mean relative  response factor of
                              a particular internal  standard  (m) relative
                              to  an  appropriate  recovery  standard,  as
                              determined from  the  five initial  calibra-
                              tion injections  (j).

      7.7.2 Criteria  for Acceptable  Calibration -  The  criteria listed
below for acceptable  calibration must be met before sample analyses are
performed.

            7.7.2.1     The percent relative standard deviations for the
      mean response factors [RFn and RFm] from  the 17 unlabeled standards
      must  not  exceed  + 20  percent,  and  those  for  the  nine  labeled
      reference compounds must not exceed + 30  percent.
                             8290  -  28                         Revision 0
                                                          September 1994

-------
            7.7.2.2     The S/N for the GC signals present in every SICP
      (including the ones for the labeled standards)  must be > 10.

            7.7.2.3     The ion abundance ratios  (Table 8) must be within
      the specified control limits.

            NOTE:       If  the  criterion  for  acceptable  calibration
                        listed  in  Sec.  7.7.2.1   is  met,  the  analyte-
                        specific RF can then be considered independent of
                        the analyte quantity for the  calibration concen-
                        tration range.  The mean  RFs will be used for all
                        calculations   until   the  routine   calibration
                        criteria (Sec.  7.7.4)  are no longer met.  At such
                        time, new mean RFs will  be calculated from a new
                        set of injections of the calibration solutions.

      7.7.3 Routine Calibration  (Continuing Calibration Check) - Routine
calibrations must be performed at the beginning of a 12-hour period after
successful  mass resolution  and GC  resolution   performance  checks.   A
routine calibration is also required at the end of a  12-hour shift.

            7.7.3.1     Inject  2 ^L  of  the  concentration  calibration
      solution HRCC-3 standard  (Table  5).   By using the  same HRGC/HRMS
      conditions as used  in Sees.  7.6.1 and 7.6.2, determine and document
      an acceptable calibration as provided in Sec. 7.7.4.

      7.7.4 Criteria  for  Acceptable Routine  Calibration  - The following
criteria must be met before further analysis is performed.

            7.7.4.1     The measured RFs  [RFn for  the unlabeled standards]
      obtained during the  routine  calibration runs  must be  within  + 20
      percent   of  the  mean   values   established  during   the  initial
      calibration (Sec.  7.7.1.4.5).

            1.1 A.2     The measured RFs  [RFm for the labeled standards]
      obtained  during the  routine calibration  runs  must  be  within
      + 30 percent of  the  mean  values  established  during  the  initial
      calibration (Sec.  7.7.1.4.7).

            7.7.4.3     The ion abundance ratios  (Table 8) must be within
      the allowed control  limits.

            7.7.4.4     If either one of the criteria  in Sees. 7.7.4.1 and
      1.1 A.2  is not satisfied, repeat one more  time.   If these criteria
      are still  not  satisfied,  the  entire  routine calibration  process
      (Sec.  7.7.1) must  be reviewed.    It is realized  that it  may not
      always be possible  to achieve  all RF criteria.  For example, it has
      occurred that the RF  criteria  for 13C12-HpCDD and 13C12-OCDD were not
      met,  however,   the   RF   values   for  the  corresponding  unlabeled
      compounds were  routinely within the  criteria  established in the
      method.   In these cases,  24 of the  26 RF parameters have met the QC
      criteria, and the  data quality  for  the unlabeled HpCDD  and  OCDD
      values were not compromised as a result of the calibration event.


                            8290 -  29                         Revision 0
                                                          September 1994

-------
      In these situations,  the analyst must  assess  the  effect on overall
      data quality as  required for the data quality objectives and decide
      on appropriate action.   Corrective  action would  be in  order,  for
      example, if the compounds  for  which  the RF criteria were  not  met
      included both the unlabeled and the corresponding  internal  standard
      compounds.   If the  ion abundance ratio criterion  (Sec.  7.7.4.3) is
      not satisfied, refer  to the note in  Sec.  7.7.1.4.2  for  resolution.

            NOTE:      An   initial   calibration  must   be  carried  out
                       whenever the  HRCC-3, the sample  fortification, or
                       the recovery standard solution is replaced by a
                       new solution  from  a  different lot.

7.8   Analysis

      7.8.1 Remove  the sample or  blank extract  (from Sec.  7.5.3.6) from
storage.   With a stream of  dry,  purified nitrogen, reduce  the extract
volume to 10 /ul_ to  50 juL.

      NOTE: A final volume  of 20 ML or  more  should  be used  whenever
            possible.  A 10  juL final volume is  difficult to  handle,  and
            injection  of  2  /uL out  of 10 /uL  leaves little  sample  for
            confirmations  and repeat injections, and for  archiving.

      7.8.2 Inject  a  2 /iL  aliquot of  the extract  into  the  GC,  operated
under the  conditions  that  have  been established to produce acceptable
results with the performance check solution (Sees.  7.6.1 and 7.6.2).

      7.8.3 Acquire SIM data  according to  Sees.  7.6.2 and 7.6.3.  Use the
same  acquisition  and  mass  spectrometer operating  conditions previously
used to determine the relative response factors  (Sees.  7.7.1.4.4 through
7.7.1.4.7).   Ions characteristic of polychlorinated diphenyl   ethers  are
included in the descriptors listed in Table 6.

      NOTE: The  acquisition period must at least encompass the PCDD/PCDF
            overall  retention time  window  previously  determined  (Sec.
            8.2.1.3). Selected ion current profiles  (SICP) for the lock-
            mass  ions (one per mass descriptor)  must  also be recorded and
            included  in  the  data package.   These  SICPs  must  be true
            representations   of  the  evolution   of  the  lock-mass  ions
            amplitudes during the  HRGC/HRMS run (see Sec.  8.2.2 for  the
            proper level  of reference compound to be  metered  into the ion
            chamber.)  The analyst may  be required to monitor a PFK ion,
            riot  as a  lock-mass,  but as a regular  ion,  in  order to meet
            this  requirement.  It is recommended to examine the lock-mass
            ion  SICP  for obvious basic  sensitivity and  stability changes
            of the instrument during the GC/MS run that could affect the
            measurements   [Tondeur  et  al.,  1984,  1987].    Report  any
            discrepancies  in  the  case narrative.
                             8290  -  30                         Revision 0
                                                          September 1994

-------
      7.8.4  Identification Criteria  -  For  a  gas  chromatographic peak to
be  identified  as  a  PCDD or  PCDF,   it  must  meet  all  of  the  following
criteria:

             7.8.4.1     Retention Times

                  7.8.4.1.1    For 2,3,7,8-substituted  congeners,  which
             have  an  isotopically labeled  internal  or  recovery standard
             present  in the sample extract  (this represents a total of 10
             congeners including OCDD; Tables  2  and  3), the retention time
             (RRT;  at maximum peak height) of  the sample components (i.e.,
             the two  ions  used  for quantitation purposes  listed in Table
             6)  must be  within  -1  to  +3  seconds  of  the  isotopically
             labeled  standard.

                  7.8.4.1.2    For 2,3,7,8-substituted compounds  that do
             not have an isotopically labeled  internal standard present in
             the sample extract (this represents a total of six congeners;
             Table  3), the retention  time must fall  within 0.005 retention
             time  units  of the relative retention  times  measured in the
             routine  calibration.  Identification of OCDF is based on its
             retention time relative  to 13C12-OCDD  as determined from the
             daily  routine calibration  results.

                  7,8.4.1.3    For non-2,3,7,8-substituted  compounds (tetra
             through  octa;  totaling   119 congeners), the  retention  time
             must  be  within  the corresponding  homologous  retention  time
             windows  established by analyzing  the column performance check
             solution (Sec. 8.1.3).

                  7.8.4.1.4    The ion current responses  for both ions used
             for quantitative purposes  (e.g.,  for TCDDs:  m/z 319.8965 and
             321.8936) must reach maximum simultaneously  (+ 2 seconds).

                  7.8.4.1.5    The ion current responses  for both ions used
             for the  labeled standards  (e.g.,  for 13C12-TCDD: m/z 331.9368
             and  m/z  333.9339)  must  reach maximum simultaneously  (+  2
             seconds).

                  NOTE:  The analyst  is  required to  verify the presence of
                        1,2,8,9-TCDD and 1,3,4,6,8-PeCDF  (Sec.  8.1.3) in
                        the   SICPs   of  the  daily  performance  checks.
                        Should  either  one  compound  be  missing,   the
                        analyst is required to  take corrective  action as
                        it may  indicate a potential  problem  with  the
                        ability to detect all the  PCDDs/PCDFs.

             7.8.4.2     Ion Abundance Ratios

                  7.8.4.2.1    The integrated  ion currents  for the two ions
             used for quantitation purposes must have a  ratio  between the
             lower  and upper limits established for the  homologous series
                            8290 - 31                         Revision  0
                                                          September 1994

-------
            to  which  the  peak  is assigned.  See  Sees.  7.7.1.4.1  and
            7.7.1.4.2  and Table 8  for details.

            7.8.4.3     Signal-to-Noise  Ratio

                  7.8.4.3.1    All  ion current intensities must  be  > 2.5
            times noise level for positive identification of a PCDD/PCDF
            compound or a group of coeluting  isomers.  Figure 6 describes
            the  procedure  to be  followed  for the determination of the
            S/N.

            7.8.4.4     Polychlorinated  Diphenyl  Ether Interferences

                  7.8.4.4.1    In  addition  to  the  above  criteria,  the
            identification of a GC peak as a PCDF can only be made  if no
            signal having a  S/N > 2.5 is  detected at the same retention
            time  (+  2 seconds)  in the  corresponding  polychlorinated
            diphenyl ether (PCDPE, Table 6)  channel.

7.9   Calculations

      7.9.1 For  gas  chromatographic peaks  that  have met  the  criteria
outlined in Sees. 7.8.4.1.1 through 7.8.4.3.1,  calculate the concentration
of the PCDD or PCDF compounds using the formula:

                  Av  n
               x  A  *

      Ax      ==     sum of the integrated ion abundances of the quantitation
                  ions  (Table  6)  for  unlabeled  PCDDs/PCDFs,

      Ajs      =     sum of the integrated ion abundances of the quantitation
                  ions  (Table  6)  for  the labeled  internal  standards,

      Qis      =     quantity,  in pg,  of the internal  standard added to the
                  sample  before extraction,

      W      =     weight,  in g, of the sample (solid or organic liquid),
                  or volume  in ml of  an  aqueous sample,  and

      RFn    =     calculated mean relative response factor  for  the analyte
                  [RFn with  n  = 1 to  17;  Sec. 7.7.1.4.5].

      If the analyte is identified as one of the 2,3,7,8-substituted PCDDs
or PCDFs, RFn is the value calculated using the equation in Sec.  7.7.1.4.5.
However,   if  it  is  a non-2,3,7,8-substituted congener, the RF(k) value is

                             8290 - 32                        Revision 0
                                                          September 1994

-------
the one calculated using the equation in Sec. 7.7.1.4.6.2.   [RFk  k = 27
to 30].

      7.9.2  Calculate the percent recovery of the nine internal standards
measured in the sample extract, using the formula:

                                              Ais   x   Qre
      Internal standard percent recovery  =  	1=-   x  100
                                            Qis x  Ars x RFm

where:

      Ais     =    sum of the integrated ion abundances of the quantitation
                  ions  (Table  6)  for the labeled  internal  standard,

      Ars     =    sum of the integrated ion abundances of the quantitation
                  ions  (Table  6)  for the labeled  recovery standard; the
                  selection  of the recovery standard depends on the type
                  of  congeners  (see Table 5,  footnotes),

      Qis     =    quantity,  in  pg,  of the internal  standard added to the
                  sample  before extraction,

      Qrs     =    quantity,  in  pg,  of the recovery standard added to the
                  cleaned-up sample residue before HRGC/HRMS analysis, and

      RFm =       calculated mean relative response factor  for the labeled
                  internal standard relative to the appropriate  (see Table
                  5,  footnotes) recovery standard.	This  represents the
                  mean obtained in Sec. 7.7.1.4.7 [RFm with m  =  18 to 26].

      NOTE:  For  human  adipose tissue,  adjust the percent recoveries by
             adding  1 percent  to  the calculated value to  compensate for
             the   1 percent  of  the  extract  diverted   for  the  lipid
             determination.

      7.9.3  If  the  concentration  in the  final  extract  of  any  of the
fifteen  2,3,7,8-substituted PCDD/PCDF  compounds  (Table  3)  exceeds the
upper method calibration limits (MCL) listed in Table 1  (e.g.,  200 pg//uL
for TCDD in soil), the linear range of response versus concentration may
have been exceeded, and a second analysis of the sample (using a one tenth
aliquot) should be undertaken.  The volumes of the internal and recovery
standard solutions should  remain the same  as described  for the sample
preparation  (Sees. 7.1  to  7.9.3).   For the other  congeners (including
OCDD), however, report the  measured  concentration  and  indicate that the
value exceeds the MCL.

             7.9.3.1     If   a   smaller   sample   size  would  not   be
      representative  of the  entire sample, one of the  following options is
      recommended:

      (1)    Re-extract an additional aliquot of sufficient  size to insure
      that  it is  representative  of the  entire sample.  Spike  it  with  a

                            8290  - 33                         Revision  0
                                                          September 1994

-------
      higher concentration of internal standard.  Prior to GC/MS analysis,
      dilute the  sample so  that  it has  a concentration  of  internal
      standard  equivalent to that  present  in the calibration  standard.
      Then,  analyze the diluted  extract.

      (2)    Re-extract an additional  aliquot of sufficient size to insure
      that it  is representative of the  entire  sample.   Spike  it  with  a
      higher concentration  of internal standard.  Immediately  following
      extraction, transfer the sample to a volumetric flask and dilute to
      known  volume.    Remove  an appropriate aliquot  and  proceed  with
      cleanup  and analysis.

      (3)    Use the original  analysis data to  quantitate the internal
      standard  recoveries.   Respike  the  original  extract (note  that no
      additional cleanup is  necessary) with  100  times the usual quantity
      of internal standards.  Dilute the re-spiked extract by a factor of
      100.  Reanalyze  the diluted  sample   using  the  internal  standard
      recoveries  calculated  from  the  initial  analysis  to  correct  the
      results  for losses during  isolation and cleanup.

      7.9.4  The total concentration for each homologous series of PCDD and
PCDF  is calculated by  summing up  the  concentrations of all  positively
identified isomers  of each homologous series. Therefore, the total should
also  include the 2,3,7,8-substituted congeners.  The total number of GC
signals included  in the homologous  total   concentration  value  must be
specified in the report.  If an isomer is not detected,  use  zero (0)  in
this calculation.

      7.9.5  Sample   Specific  Estimated   Detection   Limit  -  The  sample
specific estimated  detection limit  (EDL)  is  the  concentration of a given
analyte required to  produce a signal with a peak height  of at  least 2.5
times  the  background  signal  level.   An   EDL  is   calculated  for  each
2,3,7,8-substituted congener that is not  identified,  regardless of whether
or not other non-2,3,7,8-substituted  isomers are present.  Two methods of
calculation can be  used, as  follows, depending on  the type  of response
produced during the analysis of a particular sample.

            7.9.5.1     Samples  giving a response for  both  quantitation
      ions  (Tables  6 and 9) that  is  less than 2.5  times  the  background
      level.

                 7.9.5.1.1    Use   the  expression   for   EDL   (specific
            2,3,7,8-substituted PCDD/PCDF)  below to calculate an EDL for
            each absent 2,3,7,8-substituted  PCDD/PCDF (i.e.,  S/N < 2.5).
            The background level is determined by measuring the range of
            the noise  (peak to peak)  for the two quantitation ions (Table
            6)  of  a  particular  2,3,7,8-substituted  isomer   within  an
            homologous   series,   in   the   region   of  the   SICP  trace
            corresponding to the elution of  the internal standard  (if the
            congener possesses an internal   standard) or in the region of
            the  SICP  where  the  congener  is  expected  to  elute  by
            comparison  with  the  routine  calibration data  (for  those
            congeners   that   do  not  have  a   13C-labeled   standard),


                            8290 - 34                         Revision  0
                                                          September 1994

-------
            multiplying  that  noise  height  by  2.5,  and  relating  the
            product to an estimated concentration that would produce that
            peak  height.

                  Use  the  formula:

                                                      2.5  x   Hx   x  Qis
            EDL  (specific  2,3,7,8-subst.  PCDD/PCDF)  =
                                                       His x W x RFn
            where:
                  EDL  = estimated   detection   limit   for   homologous
                        2,3,7,8-substituted PCDDs/PCDFs.

                  Hx  =  sum of  the  height of  the  noise level  for each
                        quantitation  ion   (Table  6)  for  the  unlabeled
                        PCDDs/PCDFs, measured as shown in Figure 6.

                  His =   sum of  the  height of  the  noise level  for each
                        quantitation  ion  (Table  6)  for  the  labeled
                        internal standard,  measured  as  shown in Figure 6.

                  W,  RFn,  and  Qis retain the same meanings  as  defined in
             Sec.  7.9.1.

             7.9.5.2     Samples  characterized  by a  response  above  the
      background level with a S/N of  at  least  2.5  for both quantitation
      ions (Tables 6 and 9).

                  7.9.5.2.1    When  the response  of a  signal  having  the
             same  retention  time as  a  2,3,7,8-substituted  congener has a
             S/N  in excess  of 2.5  and does  not meet  any of  the other
             qualitative  identification criteria  listed in Sec.  7.8.4,
             calculate  the  "Estimated Maximum Possible  Concentration"
             (EMPC)  according  to the  expression  shown in  Sec. 7.9.1,
             except that Ax in Sec. 7.9.1  should represent the sum of the
             area  under the  smaller  peak  and  of the  other  peak  area
             calculated using  the theoretical chlorine  isotope  ratio.

      7.9.6  The relative percent difference (RPD) of  any duplicate sample
results are calculated as follows:

               I S, - S2 |
      RPD  =  	    x  100
             (S, + S2 ) / 2


      S.,  and  S2 represent  sample  and duplicate  sample results.

      7.9.7  The 2,3,7,8-TCDD toxicity equivalents (TE)  of PCDDs and PCDFs
present  in  the sample are  calculated,  if requested  by the  data user,
according to the method recommended by the Chlorinated Dioxins Workgroup

                            8290 -  35                         Revision 0
                                                          September 1994

-------
(CDWG) of the EPA and the Center for Disease Control  (CDC).   This method
assigns a 2,3,7,8-TCDD toxicity equivalency  factor  (TEF)  to  each of the
fifteen 2,3,7,8-substituted  PCDDs  and PCDFs  (Table  3)  and to  OCDD and
OCDF, as shown in Table  10.   The 2,3,7,8-TCDD equivalent of the PCDDs and
PCDFs present in the sample is calculated by summing the TEF  times their
concentration for each of the compounds or groups of compounds listed in
Table 10.  The exclusion of  other  homologous  series  such  as  mono-, di-,
and tri- chlorinated dibenzodioxins and dibenzofurans does not mean that
they are non-toxic.  However,  their  toxicity,  as  known  at this time, is
much lower than  the toxicity  of the compounds listed in Table  10. The
above procedure for calculating the 2,3,7,8-TCDD toxicity equivalents is
not claimed by the CDWG to be based  on a thoroughly established scientific
foundation.  The procedure, rather,  represents a "consensus recommendation
on science policy".   Since the procedure may  be  changed  in  the future,
reporting requirements  for PCDD and  PCDF  data would still  include the
reporting of the analyte  concentrations of  the  PCDD/PCDF congener as
calculated in Sees. 7.9.1 and  7.9.4.

             7.9.7.1     Two GC Column TEF Determination

                  7.9.7.1.1    The concentration of 2,3,7,8-TCDD (see note
             below), is calculated from the analysis of the sample extract
             on  the  60  m  DB-5  fused  silica  capillary  column.    The
             experimental  conditions remain  the  same as  the  conditions
             described  previously  in Sec. 7.8, and  the  calculations are
             performed  as  outlined  in  Sec.  7.9.    The  chromatographic
             separation  between the  2,3,7,8-TCDD  and its close  eluters
             (1,2,3,7/1,2,3,8-TCDD  and 1,2,3,9-TCDD)  must be equal or less
             than  25 percent  valley.

                  7.9.7.1.2    The  concentration  of  the 2,3,7,8-TCDF is
             obtained  from  the  analysis of the  sample extract on the 30 m
             DB-225  fused silica capillary  column.   However,  the GC/MS
             conditions must  be altered so that:  (1)  only the first three
             descriptors   (i.e.,  tetra-,  penta-,  and   hexachlorinated
             congeners)  of  Table 6  are used;  and  (2)  the  switching time
             between   descriptor  2   (pentachlorinated   congeners)  and
             descriptor   3   (hexachlorinated   congeners)   takes   place
             following    the    elution    of   13C12-l,2,3,7,8-PeCDD.   The
             concentration  calculations are performed as outlined in Sec.
             7.9.  The chromatographic separation between the 2,3,7,8-TCDF
             and  its close eluters (2,3,4,7-TCDF and 1,2,3,9-TCDF) must be
             equal  or  less  than 25  percent  valley.

                  NOTE:  The confirmation  and  quantitation of 2,3,7,8-TCDD
                        (Sec. 7.9.7.1.1)  may be accomplished  on  the SP-
                        2330  GC column   instead  of the  DB-5  column,
                        provided the criteria listed in  Sec.  8.2.1 are
                        met  and  the  requirements   described  in  Sec.
                        8.3.2 are followed.
                             8290  -  36                         Revision 0
                                                          September 1994

-------
                        7.9.7.1.3   For  a  gas  chromatographic  peak  to  be
                   identified as a 2,3,7,8-substituted  PCDD/PCDF congener, it
                   must  meet  the  ion  abundance  and  signal-to-noise  ratio
                   criteria listed  in  Sees.  7.8.4.2 and 7.8.4.3, respectively.
                   In   addition,  the  retention  time  identification  criterion
                   described in Sec.  7.8.4.1.1 applies  here  for congeners for
                   which a carbon-labeled analogue is  available in the sample
                   extract.  However,  the relative retention  time (RRT) of the
                   2,3,7,8-substituted  congeners  for  which  no  carbon-labeled
                   analogues are available  must fall  within 0.006 units of the
                   carbon-labeled  standard   RRT.     Experimentally,   this  is
                   accomplished by  using the attributions described in Table 11
                   and the  results from the  routine  calibration  run  on  the
                   SP-2330 column.
8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality control (QC) procedures.
Quality control to  validate  sample extraction is covered  in  Method  3500.   If
extract cleanup was performed,  follow the QC  in Method 3600 and in the specific
cleanup method.

      8.2   System  Performance  Criteria -  System  performance   criteria  are
presented below.  The laboratory may use  the  recommended GC column described in
Sec. 4.2.   It must be documented that all  applicable system performance criteria
(specified in Sees. 8.2.1 and 8.2.2) were met before analysis of any sample is
performed.  Sec. 7.6.1  provides  recommended  GC conditions  that  can  be used to
satisfy the required  criteria.   Figure  3 provides a typical  12-hour analysis
sequence,  whereby  the  response factors  and  mass  spectrometer resolving power
checks must be performed at the beginning and the  end of each  12-hour period of
operation.  A GC column performance check is only required at the beginning of
each 12-hour period  during which samples are analyzed.  An HRGC/HRMS method blank
run is required between a  calibration  run and the first  sample  run.   The same
method blank extract may thus be analyzed  more than once if the number of samples
within a batch requires more than 12 hours of analyses.

            8.2.1  GC  Column  Performance

                   8.2.1.1     Inject 2 /iL (Sec. 4.1.1) of the column performance
            check solution (Sec.  5.7)  and acquire  selected  ion monitoring (SIM)
            data as described  in Sec.  7.6.2  within  a total  cycle time  of  < 1
            second  (Sec. 7.6.3.1).

                   8.2.1.2     The chromatographic  separation  between 2,3,7,8-
            TCDD and  the  peaks representing  any  other unlabeled  TCDD isomers
            must be resolved with a valley of <  25 percent (Figure 4), where:

                   Valley  percent   =   (x/y)  (100)

                   x =   measured  as in Figure 4 from the  2,3,7,8-closest TCDD
                        eluting isomer, and
                   y =   the peak  height of 2,3,7,8-TCDD.


                                   8290 - 37                         Revision 0
                                                                September 1994

-------
       It  is the  responsibility of the  laboratory to  verify  the
conditions suitable for the appropriate resolution of 2,3,7,8-TCDD
from  all  other TCDD  isomers.   The GC  column performance  check
solution also contains the  known  first  and  last PCDD/PCDF eluters
under the conditions specified  in  this  protocol.   Their retention
times  are  used to  determine the  eight  homologue  retention  time
windows  that   are  used   for   qualitative   (Sec.  7.8.4.1)   and
quantitative purposes.  All  peaks  (that includes 13C12-2,3,7,8-TCDD)
should be labeled  and identified on the chromatograms.  Furthermore,
all first  eluters  of a  homologous  series should  be labeled with the
letter  F,   and  all last eluters of a  homologous  series  should be
labeled with  the  letter  L  (Figure 4  shows an  example of  peak
labeling for TCDD isomers).  Any  individual  selected  ion current
profile'(SICP)  (for the tetras,  this would be the SICP for m/z 322
and m/z 304) or  the reconstructed homologue  ion  current (for the
tetras, this would correspond to m/z 320 + m/z 322 + m/z 304 + m/z
306) constitutes an acceptable form of data presentation.  An SICP
for the labeled compounds (e.g., m/z 334 for labeled TCDD) is also
required.

       8.2.1.3     The retention  times  for  the switching of SIM ions
characteristic  of  one  homologous  series  to  the  next  higher
homologous series  must  be indicated in  the  SICP.  Accurate switching
at  the appropriate  times   is absolutely  necessary for accurate
monitoring of  these  compounds.   Allowable tolerance on  the daily
verification with  the GC performance check solution should be better
than  10 seconds  for  the   absolute  retention  times  of all  the
components of the mixture.   Particular caution should be exercised
for the switching  time between  the last  tetrachlorinated congener
(i.e., 1,2,8,9-TCDD) and the first  pentachlorinated congener (i.e.,
1,3,4,6,8-PeCDF),  as these  two compounds elute within 15  seconds of
each  other  on  the 60  m  DB-5 column.    A  laboratory with  a GC/MS
system that is  not capable of detecting both congeners (1,2,8,9-TCDD
and  1,3,4,6,8-PeCDF)   within  one  analysis  must  take  corrective
action.  If the recommended column is  not used, then the first and
last   eluting   isomer   of   each   homologue   must  be   determined
experimentally on  the  column which  is  used, and  the  appropriate
isomers must then  be used for window definition and switching times.

8.2.2  Mass Spectrometer  Performance

       8.2.2.1     The  mass  spectrometer  must  be  operated  in  the
electron ionization  mode.   A static  resolving power  of at least
10,000  (10 percent  valley  definition)   must  be   demonstrated  at
appropriate masses  before   any  analysis  is  performed  (Sec.  7.8).
Static resolving power  checks  must  be performed at  the beginning and
at  the  end of each 12 hour period of operation.   However,  it is
recommended  that  a  check   of the static  resolution  be  made  and
documented before and after each analysis.  Corrective action must
be  implemented whenever the resolving  power  does  not  meet  the
requirement.
                       8290 -  38                        Revision 0
                                                    September 1994

-------
            8.2.2.2     Chromatography time for PCDDs  and  PCDFs  exceeds
      the long term mass  stability of the mass spectrometer.  Because the
      instrument is operated in the  high-resolution mode,  mass drifts of
      a few ppm (e.g.,  5  ppm in mass) can have serious adverse effects on
      instrument  performance.  Therefore,  a  mass  drift  correction  is
      mandatory.  To  that effect, it is recommended to select a lock-mass
      ion from the  reference compound (PFK is recommended) used for tuning
      the  mass  spectrometer.   The  selection  of  the  lock-mass   ion  is
      dependent  on  the  masses  of  the  ions  monitored  within  each
      descriptor.  Table  6 offers some  suggestions for the lock-mass ions.
      However,  an  acceptable  lock-mass   ion  at  any  mass  between  the
      lightest and heaviest  ion  in each descriptor  can be used to  monitor
      and correct mass  drifts.  The level  of the reference compound (PFK)
      metered into the  ion  chamber  during HRGC/HRMS analyses  should be
      adjusted so that the  amplitude of  the most  intense  selected lock-
      mass  ion  signal  (regardless  of  the descriptor  number) does  not
      exceed 10 percent  of the full  scale deflection for  a  given set of
      detector parameters.   Under those  conditions,  sensitivity  changes
      that  might  occur  during  the  analysis  can  be  more  effectively
      monitored.

      NOTE: Excessive  PFK (or any other reference substance)  may cause
            noise  problems  and contamination of the ion source resulting
            in  an  increase  in downtime for source  cleaning.

            8.2.2.3     Documentation  of the  instrument  resolving power
      must then be  accomplished by recording the peak profile of the high-
      mass reference  signal  (m/z  380.9760) obtained during the above peak
      matching experiment by using the low-mass PFK  ion  at m/z 304.9824 as
      a  reference.    The minimum  resolving  power  of 10,000  must  be
      demonstrated on the high-mass ion while it is  transmitted at a lower
      accelerating voltage  than the  low-mass  reference  ion, which  is
      transmitted at full sensitivity.   The  format of the  peak  profile
      representation  (Figure 5)  must  allow manual  determination  of  the
      resolution,  i.e.,  the horizontal  axis  must  be a calibrated  mass
      scale (amu  or  ppm  per division).   The  result  of  the  peak width
      measurement  (performed  at  5  percent   of  the  maximum,  which
      corresponds to  the  10  percent valley definition) must appear on the
      hard copy and cannot exceed 100 ppm  at m/z 380.9760 (or 0.038 amu at
      that particular mass).

8.3   Quality Control Samples

      8.3.1 Performance  Evaluation  Samples  -  Included  among the  samples
in all  batches  may be  samples (blind or double blind) containing known
amounts of  unlabeled 2,3,7,8-substituted  PCDDs/PCDFs or  other PCDD/PCDF
congeners.

      8.3.2 Performance  Check Solutions

            8.3.2.1     At the beginning  of  each  12-hour period during
      which samples are  to  be  analyzed,  an  aliquot of the  1)  GC column
      performance  check  solution and  2) high-resolution  concentration
      calibration solution No.  3 (HRCC-3; see  Table 5)  shall  be analyzed

                             8290 - 39                         Revision 0
                                                          September 1994

-------
to demonstrate  adequate GC  resolution  and  sensitivity,  response
factor reproducibility,  and mass range calibration,  and to establish
the PCDD/PCDF retention  time windows.  A  mass  resolution check shall
also be performed to demonstrate adequate mass resolution using an
appropriate  reference  compound  (PFK  is  recommended).    If  the
required criteria are not met, remedial  action must be taken before
any samples are analyzed.

       8.3.2.2    To validate positive sample data, the routine or
continuing calibration  (HRCC-3;  Table 5)  and the  mass  resolution
check  must  be performed also  at the end  of each  12-hour period
during which samples  are analyzed.  Furthermore, an HRGC/HRMS method
blank run must be recorded following a calibration run and the first
sample run.

            8.3.2.2.1   If the laboratory operates  only during one
       period  (shift)   each  day of   12  hours  or  less,  the  GC
       performance check solution must be analyzed only once (at the
       beginning of the period) to validate  the data acquired during
       the period.    However, the  mass  resolution  and  continuing
       calibration checks must be performed  at the beginning as well
       as at  the  end  of  the  period.

            8.3.2.2.2   If   the    laboratory    operates   during
       consecutive  12-hour  periods  (shifts), analysis  of the  GC
       performance check solution must be performed  at the beginning
       of each 12-hour period.   The mass resolution and continuing
       calibration checks  from the  previous period  can be used for
       the beginning  of  the  next  period.

       8.3.2.3    Results of at least one analysis  of the GC column
performance check solution and of two mass resolution and continuing
calibration checks must be reported with the sample data collected
during a  12 hour period.

       8.3.2.4    Deviations  from  criteria  specified  for the  GC
performance check or  for the mass  resolution check invalidate all
positive  sample data collected between analyses of the performance
check  solution, and the extracts from those positive samples shall
be reanalyzed.

       If the routine calibration run fails  at the beginning of a 12
hour shift, the instructions in Sec. 7.7.4.4 must be followed.  If
the continuing calibration check performed at the end of a 12 hour
period fails  by  no  more than 25 percent RPD for the  17 unlabeled
compounds and 35 percent RPD for  the  9 labeled reference compounds,
use the  mean RFs from  the  two daily routine calibration  runs to
compute the analyte concentrations, instead of the RFs obtained from
the initial  calibration.   A new initial calibration  (new  RFs) is
required  immediately  (within two hours)  following  the analysis of
the  samples,  whenever  the  RPD  from  the  end-of-shift  routine
calibration exceeds  25 percent or 35 percent,  respectively.  Failure
to  perform a  new  initial   calibration  immediately following  the
analysis of the  samples will  automatically require reanalysis of all

                       8290  - 40                         Revision 0
                                                    September 1994

-------
      positive sample extracts  analyzed  before the  failed  end-of-shift
      continuing calibration check.

      8.3.3 The  GC  column  performance  check mixture,  high-resolution
concentration calibration   solutions,   and  the   sample   fortification
solutions may be obtained from  the  EMSL-CIN.   However,  if not available
from  the  EMSL-CIN,  standards can  be obtained  from other  sources,  and
solutions  can  be  prepared  in  the  laboratory.   Concentrations of  all
solutions  containing  2,3,7,8-substituted  PCDDs/PCDFs,  which  are  not
obtained from the  EMSL-CIN, must be  verified  by  comparison  with the  EPA
standard solutions that are available from the EMSL-CIN.

      8.3.4 Field  Blanks - Each  batch of  samples usually contains a field
blank sample  of uncontaminated soil,  sediment or  water  that  is  to  be
fortified before analysis according  to Sec. 8.3.4.1.  In addition to this
field blank, a batch of  samples  may  include a rinsate, which is a portion
of the solvent  (usually trichloroethylene) that was used to rinse sampling
equipment.  The rinsate  is  analyzed  to assure  that the  samples  were  not
contaminated by the sampling equipment.

            8.3.4.1     Fortified  Field  Blank

                 8.3.4.1.1    Weigh  a 10 g portion or use 1 L (for aqueous
            samples) of  the specified field  blank sample  and  add  100  juL
            of  the  solution containing  the  nine  internal   standards
            (Table 2) diluted with  1.0 mL  acetone  (Sec. 7.1).

                 8.3.4.1.2    Extract by using the procedures beginning
            in  Sees. 7.4.5 or  7.4.6,  as applicable, add  10  juL of  the
            recovery standard solution (Sec.  7.5.3.6) and  analyze  a 2  /zL
            aliquot  of the  concentrated  extract.

                 8.3.4.1.3    Calculate the concentration  (Sec. 7.9.1)  of
            2,3,7,8-substituted PCDDs/PCDFs  and  the percent recovery  of
            the internal standards  (Sec. 7.9.2).

                 8.3.4.1.4    Extract  and   analyze  a   new   simulated
            fortified  field  blank   whenever  new  lots  of  solvents  or
            reagents  are  used   for  sample  extraction  or  for  column
            chromatographic procedures.

            8.3.4.2     Rinsate  Sample

                 8.3.4.2.1    The  rinsate sample must be  fortified like
            a regular sample.

                 8.3.4.2.2    Take  a  100  mL  (+ 0.5  mL)  portion of  the
            sampling equipment rinse solvent  (rinsate sample), filter,  if
            necessary, and add 100 /uL of  the solution containing the nine
            internal standards  (Table 2).

                 8.3.4.2.3    Using   a  KD  apparatus,  concentrate   to
            approximately 5 mL.
                            8290 - 41                         Revision  0
                                                          September 1994

-------
            NOTE:  As an option, a rotary evaporator may be used in
                  place of the KD  apparatus  for the concentration
                  of the rinsate.

            8.3.4.2.4    Transfer the 5 ml  concentrate  from the KD
       concentrator  tube  in  1  mL  portions  to a  1 ml  minivial,
       reducing  the volume  in the minivial   as necessary  with  a
       gentle  stream of dry nitrogen.

            8.3.4.2.5    Rinse  the  KD concentrator  tube  with  two
       0.5  ml  portions  of hexane and transfer  the rinses to the 1 ml
       minivial.   Blow  down with  dry  nitrogen  as necessary.

            8.3.4.2.6    Just before  analysis,  add  10 jitL  recovery
       standard  solution  (Table  2) and  reduce  the volume to  its
       final   volume,   as   necessary   (Sec.  7.8.1).    No  column
       chromatography is required.

            8.3.4.2.7    Analyze  an  aliquot   following  the  same
       procedures  used  to  analyze samples.

            8.3.4.2.8    Report  percent   recovery  of the   internal
       standard  and  the presence of any PCDD/PCDF compounds in jug/L
       of rinsate  solvent.

8.3.5  Duplicate Analyses

       8.3.5.1    In each batch   of  samples,  locate  the sample
specified for duplicate analysis, and analyze  a  second 10  g soil or
sediment sample portion or  1  L water  sample,  or  an  appropriate
amount of the type of matrix under consideration.

            8.3.5.1.1    The  results  of  the  laboratory  duplicates
       (percent  recovery and  concentrations of  2,3,7,8-substituted
       PCDD/PCDF compounds) should  agree  within  25 percent  relative
       difference  (difference expressed as  percentage of the mean).
       Report  all  results.

            8.3.5.1.2    Recommended actions to help locate  problems:

                  8.3.5.1.2.1  Verify     satisfactory    instrument
            performance (Sees.  8.2  and 8.3).

                  8.3.5.1.2.2  If possible, verify  that no  error was
            made while  weighing the sample portions.

                  8.3.5.1.2.3  Review the analytical  procedures with
            the  performing laboratory personnel.

8.3.6  Matrix  Spike  and Matrix  Spike  Duplicate

       8..3.6.1    Locate the sample for the MS and  MSD analyses (the
sample may be labeled  "double  volume").
                       8290 -  42                        Revision 0
                                                    September 1994

-------
                   8.3.6.2    Add  an  appropriate volume  of the  matrix spike
            fortification solution (Sec. 5.10) and of the sample fortification
            solution (Sec. 5.8),  adjusting  the fortification level as specified
            in Table 1 under IS Spiking Levels.

                   8.3.6.3    Analyze the  MS  and MSD samples  as  described in
            Sec. 7.

                   8.3.6.4    The results obtained from the MS and MSD samples
            (concentrations  of  2,3,7,8-substituted  PCDDs/PCDFs)  should agree
            within 20 percent relative difference.

      8.4   Percent Recovery of the Internal Standards -  For each sample, method
blank and  rinsate,  calculate the percent recovery  (Sec.  7.9.2).   The percent
recovery should be between 40 percent and 135 percent  for all 2,3,7,8-substituted
internal standards.

      NOTE;        A low or high  percent recovery for a blank does not require
                   discarding the  analytical  data  but   it  may  indicate  a
                   potential  problem with future  analytical  data.

      8.5   Identification Criteria

            8.5.1  If either one of  the identification criteria  appearing in
      Sees. 7.8.4.1.1 through 7.8.4.1.4 is  not met for an homologous series, it
      is reported that the sample does not contain unlabeled 2,3,7,8-substituted
      PCDD/PCDF isomers for that homologous series at the calculated detection
      limit (Sec. 7.9.5)

            8.5.2  If the  first initial identification criteria  (Sees. 7.8.4.1.1
      through 7.8.4.1.4) are met, but  the criteria appearing in Sees. 7.8.4.1.5
      and 7.8.4.2.1 are not met,  that  sample is presumed to contain interfering
      contaminants.  This must be noted on the analytical  report form, and the
      sample should be rerun or  the extract reanalyzed.

      8.6   Unused portions of samples and sample extracts should be preserved
for six months after sample receipt to allow further analyses.

      8.7   Reuse  of  glassware  is  to  be minimized  to  avoid  the  risk  of
contamination.
9.0   METHOD PERFORMANCE

      9.1   Data are currently not available.


10.0  REFERENCES

1.    "Control  of  Interferences  in the Analysis  of Human Adipose  Tissue  for
      2,3,7,8-Tetrachlorodibenzo-p-dioxin".  D.  G.  Patterson, J.S. Holler, D.F.
      Grote,  L.R.  Alexander,  C.R.  Lapeza,  R.C.  O'Connor  and  J.A.  Liddle.
      Environ. Toxicol. Chem. 5,  355-360 (1986).


                                   8290  - 43                         Revision 0
                                                                September 1994

-------
2.    "Method 8290: Analytical Procedures and Quality Assurance for Multimedia
      Analysis of Polychlorinated Dibenzo-p-Dioxins and Dibenzofurans by High-
      Resolution  Gas  Chromatography/High-Resolution  Mass  Spectrometry".   Y.
      Tondeur  and  W.F.  Beckert.    U.S.  Environmental  Protection  Agency,
      Environmental Monitoring Systems Laboratory, Las Vegas, NV.

3.    "Carcinogens - Working with Carcinogens", Department of Health, Education,
      and Welfare, Public Health Service,  Center  for Disease Control.  National
      Institute  for Occupational Safety  and  Health.  Publication  No.  77-206,
      August 1977.

4.    "OSHA  Safety and  Health  Standards, General  Industry",  (29  CFR 1910),
      Occupational Safety and Health Administration, OSHA 2206 (revised January
      1976).

5.    "Safety  in  Academic Chemistry Laboratories", American  Chemical  Society
      Publication, Committee on Chemical  Safety  (3rd  Edition, 1979.)

6.    "Hybrid  HRGC/MS/MS Method for the  Characterization  of Tetrachlorinated
      Dibenzo-p-dioxins  in Environmental  Samples." Y. Tondeur, W.J. Niederhut,
      S.R. Missler, and  J.E. Campana, Mass Spectrom.  14, 449-456 (1987).

7.    USEPA National Dioxin Study - Phase II,  "Analytical  Procedures  and Quality
      Assurance Plan  for the  Determination  of PCDD/PCDF in Fish",   EPA-Duluth,
      October 26,  1987.
11.0  SAFETY

      11.1  The following  safety  practices are excerpts from  EPA Method 613,
Sec. 4 (July 1982 version)  and amended for  use  in conjunction with this method.
The  2,3,7,8-TCDD  isomer has  been  found  to be  acnegenic,  carcinogenic,  and
teratogenic in  laboratory  animal  studies.   Other PCDDs and  PCDFs containing
chlorine atoms in positions 2,3,7,8 are known to have toxicities comparable to
that of 2,3,7,8-TCDD.   The analyst should note that  finely  divided dry soils
contaminated with  PCDDs and  PCDFs  are particularly hazardous  because  of the
potential  for inhalation and ingestion.  It is  recommended  that  such samples be
processed in a confined environment, such  as  a  hood or a glove box.  Laboratory
personnel  handling these types of samples should wear masks  fitted with charcoal
filters to prevent inhalation of dust.

      11.2  The toxicity or carcinogenicity of  each  reagent used in this method
is not precisely defined; however, each chemical compound  should be treated as
a potential  health hazard.  From this viewpoint, exposure to  these chemicals must
be kept to a minimum.  The laboratory  is responsible for maintaining a current
awareness file of OSHA regulations regarding  the safe handling of the chemicals
specified in this method.  A reference  file  of material safety  data  sheets should
be made available to all personnel  involved in  the chemical analysis of samples
suspected to contain  PCDDs  and/or  PCDFs.   Additional  references to laboratory
safety are given in references 3,  4 and 5.

      11.3  Each laboratory must develop a strict safety program for the handling
of PCDDs and PCDFs.  The laboratory practices listed below are  recommended.


                                   8290 -  44                         Revision 0
                                                                September 1994

-------
            11.3.1      Contamination  of the  laboratory  will  be minimized  by
      conducting most of the manipulations in a hood.

            11.3.2      The   effluents  of   sample   splitters  for   the   gas
      chromatograph  and roughing pumps  on  the  HRGC/HRMS  system  should  pass
      through either a column of activated charcoal  or be  bubbled through a trap
      containing oil or high boiling alcohols.

            11.3.3      Liquid waste should be dissolved in methanol or ethanol
      and irradiated with ultraviolet  light at a wavelength  less than 290 nm for
      several days  (use  F 40 BL  lamps,  or equivalent).   Using this analytical
      method, analyze the irradiated liquid wastes and dispose of the solutions
      when 2,3,7,8-TCDD and -TCDF congeners can no longer be detected.

      11.4  The following precautions were issued by Dow Chemical U.S.A. (revised
11/78) for safe handling of 2,3,7,8-TCDD in  the laboratory and amended for use
in conjunction with this method.

            11.4.1      The following statements on safe handling are as complete
      as possible  on the basis  of  available toxicological information.   The
      precautions for safe handling and  use  are necessarily general  in nature
      since  detailed,  specific  recommendations  can  be  made  only  for  the
      particular exposure and circumstances of each individual  use.   Assistance
      in evaluating  the health  hazards of particular plant conditions  may  be
      obtained from certain consulting laboratories and from State  Departments
      of Health or  of Labor, many of  which have an  industrial health service.
      The 2,3,7,8-TCDD isomer is  extremely toxic to certain kinds of laboratory
      animals.   However,  it  has  been handled  for  years  without  injury  in
      analytical and biological  laboratories.   Many techniques used in handling
      radioactive and infectious materials are applicable to 2,3,7,8-TCDD.

                   11.4.1.1    Protective Equipment:  Throw away plastic gloves,
            apron or lab coat,  safety  glasses and laboratory hood adequate for
            radioactive  work.   However, PVC  gloves  should not  be  used.

                   11.4.1.2    Training:  Workers must be  trained  in the proper
            method  of  removing  contaminated  gloves  and clothing  without
            contacting the exterior surfaces.

                   11.4.1.3    Personal Hygiene:  Thorough washing of hands and
            forearms after each manipulation  and before breaks (coffee, lunch,
            and shift).

                   11.4.1.4    Confinement:    Isolated work  area,   posted  with
            signs,  segregated  glassware and  tools,  plastic  backed  absorbent
            paper on benchtops.

                   11.4.1.5    Waste:     Good   technique   includes   minimizing
            contaminated waste.   Plastic bag  liners  should  be used  in  waste
            cans.

                   11.4.1.6    Disposal  of  Hazardous  Wastes:    Refer to  the
            November 7,  1986 issue of  the Federal  Register on  Land  Ban Rulings
            for details  concerning the handling of dioxin containing wastes.

                                  8290  - 45                         Revision  0
                                                                September 1994

-------
       11.4.1.7    Decontamination:  Personnel  -  apply a mild soap
with plenty of  scrubbing  action.   Glassware,  tools and surfaces -
Chlorothene NU  Solvent  (Trademark  of  the  Dow  Chemical Company)  is
the  least toxic  solvent  shown to  be  effective.   Satisfactory
cleaning  may  be  accomplished  by  rinsing with  Chlorothene,  then
washing with a .detergent and water.  Dish water may be disposed to
the sewer after percolation through a  charcoal  bed filter.   It is
prudent to  minimize solvent wastes  because they  require special
disposal through commercial services that are expensive.

       11.4.1.8    Laundry:  Clothing known to be contaminated should
be  disposed  with  the  precautions described  under  "Disposal  of
Hazardous  Wastes".  Laboratory  coats   or  other  clothing worn  in
2,3,7,8-TCDD  work  area  may  be laundered.    Clothing  should  be
collected in plastic bags.   Persons who convey  the  bags and launder
the clothing should be advised of the hazard and trained  in proper
handling.  The clothing may be  put  into a  washer without contact if
the launderer knows the problem.  The washer should be run through
one full cycle before being used again for other clothing.

       11.4.1.9    Wipe  Tests:    A useful  method  for determining
cleanliness of work surfaces and tools is to wipe the surface with
a piece of filter paper,  extract  the  filter paper and analyze the
extract.

       NOTE:       A   procedure   for   the   collection,    handling,
                  analysis,  and  reporting  requirements  of  wipe
                  tests   performed  within   the   laboratory   is
                  described  in Attachment  A.    The   results  and
                  decision  making  processes  are  based  on  the
                  presence of 2,3,7,8-substituted PCDDs/PCDFs.

       11.4.1.10  Inhalation:   Any  procedure  that  may  generate
airborne contamination must be  carried out  with good ventilation.
Gross losses to a ventilation  system must  not be  allowed.  Handling
of the dilute solutions normally used  in analytical  and animal work
presents  no  significant inhalation hazards  except in  case  of an
accident.

       11.4.1.11  Accidents:       Remove   contaminated   clothing
immediately, taking  precautions not  to contaminate  skin or other
articles.  Wash  exposed skin vigorously and repeatedly  until medical
attention is obtained.
                       8290 -  46                        Revision 0
                                                    September 1994

-------
                                 Attachment A

            PROCEDURES  FOR THE COLLECTION,  HANDLING, ANALYSIS,  AND
            REPORTING OF  WIPE TESTS  PERFORMED  WITHIN THE  LABORATORY

      This  procedure is designed for the periodic evaluation of potential con-
tamination by 2,3,7,8-substituted PCDD/PCDF congeners of the working areas inside
the laboratory.

      A.I   Perform the wipe  tests  on  surface  areas of  two inches  by one foot
with glass fiber paper  saturated with distilled in glass  acetone using a pair of
clean stainless steel forceps.  Use  one wiper  for each of the designated areas.
Combine the wipers to one composite sample in an extraction jar containing 200
mL distilled in glass acetone.  Place an equal  number of  unused  wipers in 200 mL
acetone and use this as  a  control.   Add  100 fj.1 of the  sample fortification
solution to each jar containing used or unused wipers  (Sec. 5.8).

            A. 1.1  Close  the  jar  containing  the  wipers  and  the  acetone  and
      extract for 20 minutes using a wrist  action shaker.  Transfer the extract
      into  a  KD apparatus fitted with a  concentration  tube and a  three ball
      Snyder  column.   Add  two  Teflon™  or  Carborundum™ boiling  chips  and
      concentrate the extract to an apparent volume of 1.0 mL on a steam bath.
      Rinse the Snyder column and  the KD  assembly  with two 1  mL  portions of
      hexane  into the  concentrator  tube,  and  concentrate its  contents  to near
      dryness  with  a gentle  stream of nitrogen.   Add  1.0  mL hexane  to  the
      concentrator tube and  swirl the  solvent on the walls.

            A.1.2  Prepare a  neutral alumina column as described in Sec.  7.5.2.2
      and follow the steps outlined in Sees. 7.5.2.3 through 7.5.2.5.
            A. 1.3  Add  10  juL of the recovery standard  solution  as described in
      Sec.  7.5.3.6.

      A.2   Concentrate the  contents  of  the vial to a  final  volume of  10  /xL
(either in  a  minivial  or in a capillary tube).   Inject 2 jtiL  of  each  extract
(wipe and control) onto a capillary column and analyze  for 2,3,7,8-substituted
PCDDs/PCDFs  as  specified  in  the  analytical  method  in  Sec.  7.8.    Perform
calculations according to Sec. 7.9.

      A.3   Report the presence  of 2,3,7,8-substituted  PCDDs  and PCDFs as  a
quantity (pg or ng) per wipe test experiment (WTE).   Under the  conditions out-
lined in this analytical  protocol, a lower  limit of calibration of 10 pg/WTE is
expected for  2,3,7,8-TCDD.    A  positive  response for  the blank  (control)  is
defined as  a  signal  in the  TCDD  retention time window  at any of  the  masses
monitored which  is equivalent  to  or above  3 pg of 2,3,7,8-TCDD  per  WTE.   For
other congeners, use  the  multiplication factors listed in Table 1, footnote (a)
(e.g., for OCDD, the lower MCL is 10 x 5 = 50 pg/WTE  and the positive response
for the blank would  be 3 x  5  =  15 pg).   Also,  report  the recoveries  of  the
internal standards during the simplified cleanup procedure.

      A.4   At a minimum,  wipe tests should  be performed when there is evidence
of contamination in the method blanks.
                                  8290  - 47                         Revision 0
                                                                September 1994

-------
      A.5   An upper limit of 25 pg per TCDD isomer and per wipe test experiment
is allowed (use multiplication factors listed in footnote (a) from Table 1 for
other congeners).  This value corresponds to 2? times the  lower calibration limit
of the  analytical  method.  Steps  to  correct the contamination must  be taken
whenever these levels are  exceeded.   To  that effect,  first  vacuum the working
places  (hoods,  benches,  sink)  using  a  vacuum cleaner  equipped  with  a  high
efficiency particulate absorbent (HEPA) filter and then  wash with a detergent.
A new set of wipes  should  be analyzed before anyone  is  allowed  to work in the
dioxin area of the laboratory after corrective action has been taken.
                                   8290 - 48                        Revision 0
                                                                September 1994

-------
                   Figure 1.
  8
              Dibenzodioxin
  8
               Dibenzofuran
General structures of dibenzo-p-dioxin and dibenzofuran.
                   8290 -  49
   Revision 0
September 1994

-------
                                  Figure 2.
                                                 M/AM
                                                 5,600
                                                 5,600
                                                  8,550
  Peak profile displays demonstrating the effect  of  the detector zero on the
measured resolving  power.   In this example,  the true resolving power is 5,600.

      A)    The  zero  was  set  too  high;  no effect  is  observed  upon  the
      measurement of the  resolving power.

      B)    The zero was  adjusted properly.

      C)    The zero was  set too low; this results in overestimating the actual
      resolving  power  because  the peak-to-peak  noise  cannot  be  measured
      accurately.
                                  8290  -  50
    Revision  0
September 1994

-------
                             Figure 3.
                    Analytical  Procedure
   8:00 AM
Mass Resolution
 Mass Accuracy
                       Thaw Sample Extract
                                I
                        Concentrate to 10
                                I
 9:00 AM
 Initial or
 Routine
Calibration
              QC Column
              Performance
11:00 AM
        Method
         Blank
8:00 PM
                   Mass
                 Resolution
         Routine
        Calibration
              Typical 12 hour analysis sequence of events.
                            8290 - 51
                              Revision 0
                           September 1994

-------
                                      Figure 4.
00
 >
_*-•

 c
 *-»
_c

 0>


JH
 0)
CC
      oo
      «n
—J—,—
 24:00
    Time
                  22:30
                                                       25:30
-,	,

   27:00
     Selected  ion  current  profile  for m/z 322  (TCDDs) produced by MS analysis of
   the GC performance check solution on a 60 m DB-5 fused silica capillary column
   under the conditions listed in Sec.  7.6.
                                      8290  -  52
                                                                       Revision 0
                                                                   September 1994

-------
                                 Figure 5.
       80-i
       60-
       40-
       20-
                                         Ref. mass 304.9824 Peak top
                                         Span. 200 ppm
System file name

Data file name

Resolution

Group number

lonization mode

Switching

Ref. masses
YVES150

A:85Z567

   10000

        1

      EI +

VOLTAGE

304.9824

380.9260
                                             M/AM—10.500
                                         Channel B 380.9260 Lock mass
                                         Span 200 ppm
 Peak profiles representing two PFK reference ions at m/z 305  and 381.  The
resolution of  the high-mass signal  is 95 ppm at  5  percent of the peak height;
  this corresponds  to a resolving power M/dM of 10,500 (10 percent valley
                               definition).
                                8290 - 53
                       Revision 0
                    September 1994

-------
                                 Figure 6.
20:00
22:00
24:00
26:00
28:00
30:00
                        Manual  determination  of S/N.

     The peak height (S) is measured between the mean noise  (lines  C  and D).
     These mean  signal  values  are obtained  by  tracing  the line between  the
     baseline average noise extremes,  El  and  E2,  and  between  the  apex  average
     noise extremes, E3 and E4, at the apex of the signal.
     NOTE:
    It is  imperative that  the instrument  interface  amplifier
    electronic zero offset be  set  high  enough  so that negative
    going baseline noise is recorded.
                                 8290 - 54
                                                     Revision 0
                                                 September 1994

-------
                                   Table 1.

            Types of Matrices, Sample Sizes and 2,3,7,8-TCDD-Based
                Method Calibration Limits (Parts per Trillion)

Lower MCLa
Upper MCLa
Weight (g)
IS Spiking
Levels (ppt)
Final Extr.
Vol. (jLtL)d
Water
0.
2
1000
1
10-50
Soil
Sediment
Paper Pulpb
01 1.0
200
10
100
10-50
Fly
Ash
1.0
200
10
100
50
Fish
Tissue
1.0
200
20
100
10-50
Human
Adipose
0 Tissue
1.0
200
10
100
10-50
Sludges,
Fuel Oil
5.0
1000
2
500
50
Still-
Bottom
10
2000
1
1000
50
a     For other congeners multiply the values  by  1 for TCDF/PeCDD/PeCDF, by 2.5
      for HxCDD/HxCDF/HpCDD/HpCDF, and by 5 for OCDD/OCDF.

b     Sample dewatered according to Sec. 6.5.

c     One half of the extract from the 20 g  sample is used for determination of
      lipid content  (Sec. 7.2.2).

d     See Sec. 7.8.1, Note.


NOTE: Chemical  reactor  residues  are  treated  as  still  bottoms  if  their
      appearances so suggest.
                                   8290  -  55
    Revision 0
September 1994

-------
                                  Table  2.

                    Composition of the Sample Fortification
                        and  Recovery  Standard Solutions8
Analyte
                       Sample Fortification
                       Solution
                       Concentration
                       (pg//LtL; Solvent:
                       Nonane)
                   Recovery Standard
                   Solution
                   Concentration
                   (pg//xL; Solvent:
                   Nonane)
13
  C12-2,3,7,8-TCDD
13C12-2,3,7,8-TCDF
13C12-1,2,3,4-TCDD
13

13
13
13

13
C12-l,2,3,7,8-PeCDD
C12-l,2,3,7,8-PeCDF

C12-l,2,3,6,7,8-HxCDD
C12-l,2,3,4,7,8-HxCDF
C12-l,2,3,7,8,9-HxCDD
13C12-l,2,3,4,6,7,8-HpCDD
13C12-l,2,3,4,6,7,8-HpCDF
13C12-OCDD
                             10
                             10
10
10

25
25
                             25
                             25
                             50
                                                     50
                                                      50
(a)  These solutions should be made  freshly every day because of the possibility
of adsorptive losses to glassware.   If these  solutions  are to be kept for more
than one  day,  then the sample fortification  solution concentrations should be
increased ten fold, and the recovery standard solution concentrations should be
doubled.  Corresponding adjustments of the spiking  volumes must then be made.
                                   8290 - 56
                                                                    Revision  0
                                                               September  1994

-------
                                   Table 3.

           The  Fifteen  2,3,7,8-Substituted  PCDD and PCDF Congeners


      PCDD                             PCDF


   2,3,7,8-TCDD(*)                   2,3,7,8-TCDF(*)

   l,2,3,7,8-PeCDD(*)                l,2,3,7,8-PeCDF(*)

   l,2,3,6,7,8-HxCDD(*)              2,3,4,7,8-PeCDF

   1,2,3,4,7,8-HxCDD                 1,2,3,6,7,8-HxCDF

   l,2,3,7,8,9-HxCDD(+)              1,2,3,7,8,9-HxCDF

   l,2,3,4,6,7,8-HpCDD(*)            l,2,3,4,7,8-HxCDF(*)

                                    2,3,4,6,7,8-HxCDF

                                    l,2,3,4,6,7,8-HpCDF(*)

                                    1,2,3,4,7,8,9-HpCDF



(*) The  13C-labeled analogue is used as an internal  standard.

(+) The  13C-labeled analogue is used as a recovery standard.
                                  8290 - 57                         Revision 0
                                                                September 1994

-------
                   Table 4.

Isomers of Chlorinated Dioxins and Furans as  a
   Function of the Number of Chlorine Atoms
Number of
Chlorine
Atoms
1
2
3
4
5
6
7
8
Total
Number of
Dioxin
Isomers
2
10
14
22
14
10
2
1
75
Number of
2,3,7,8
Isomers
—
—
—
1
1
3
1
1
7
Number of
Furan
Isomers
4
16
28
38
28
16
4
1
135
Number of
2,3,7,8
Isomers
—
—
—
1
2
4
2
1
10
                  8290  -  58
    Revision 0
September 1994

-------
                                   Table 5.

              High-Resolution Concentration Calibration Solutions
                                          Concentration (pq/uL, in Nonane)
Compound
HRCC
Unlabeled Analytes
2,3,7,8-TCDD
2,3,7,8-TCDF
1,2,3,7,8-PeCDD
1,2,3,7,8-PeCDF
2,3,
1,2,
1,2,
1,2,
1,2,
1,2,
1,2,
2,3,
1,2,
1,2,
1,2,
OCDD
OCDF
4,7,
3,4,
3,6,
3,7,
3,4,
3,6,
3,7,
4,6,
3,4,
3,4,
3,4,


8-PeCDF
7
7
,
,
8,
7
7
8
7
6
6
7


Internal
13C
13f12
u12
13C1?
13C
13/- 12
13p12
73C12
,, 12
13p
, L12
C
-2,3
-2,3
-1,2
-1,2
-1,2
-1,2
-1,2
-1,2

,
,
?
}
,
,
,
,
,
,
,
,
,
,


8-HxCDD
8-HxCDD
9-HxCDD
8-HxCDF
8-HxCDF
9-HxCDF
8-HxCDF
7,8-HpCDD
7,8-HpCDF
8,9-HpCDF


200
200
500
500
500
500
500
500
500
500
500
500
500
500
500
1,000
1,000
50
50
125
125
125
125
125
125
125
125
125
125
125
125
125
250
250
10
10
25
25
25
25
25
25
25
25
25
25
25
25
25
50
50
2
2
6
6
6
6
6
6
6
6
6
6
6
6
6
12
12
.5
.5
.25
.25
.25
.25
.25
.25
.25
.25
.25
.25
.25
.25
.25
.5
.5
1
1
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
2.
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5


Standards
7
7
3
3
3
3
3
3
,8-TCDD
,8-TCDF
,7,8-PeCDD
,7,8-PeCDF
,6,7,8-HxCDD
,4,7,8-HxCDF
,4,6,7,8-HpCDD
,4,6,7,8-HpCDF
-OCDD
Recovery
13f
V/12
13c
-1,2
-1,2

'
50
50
50
50
125
125
125
125
250
50
50
50
50
125
125
125
125
250
50
50
50
50
125
125
125
125
250
50
50
50
50
125
125
125
125
250









50
50
50
50
125
125
125
125
250









Standards
3
3
,4-TCDD(a|
,7,8,9-HxCDD(b|
50
125
50
125
50
125
50
125


50
125


(a)
    Used for recovery  determinations  of TCDD, TCDF,  PeCDD  and PeCDF internal
    standards.
(bl   Used for recovery  determinations of  HxCDD,  HxCDF, HpCDD,  HpCDF  and OCDD
    internal  standards.
                                   8290  -  59
                                            Revision 0
                                        September 1994

-------
                      Table 6.

Ions Monitored for HRGC/HRMS Analysis  of PCDDs/PCDFs
Descriptor
1









2









3









Accurate1"1
Mass
303.9016
305.8987
315.9419
317.9389
319.8965
321.8936
331.9368
333.9338
375.8364
[354.9792]
339.8597
341.8567
351.9000
353.8970
355.8546
357.8516
367.8949
369.8919
409.7974
[354.9792]
373.8208
375.8178
383.8639
385.8610
389.8156
391.8127
401.8559
403.8529
445.7555
[430.9728]
Ion
ID
M
M+2
M
M+2
M
M+2
M
M+2
M+2
LOCK
M+2
M+4
M+2
M+4
M+2
M+4
M+2
M+4
M+2
LOCK
M+2
M+4
M
M+2
M+2
M+4
M+2
M+4
M+4
LOCK
Elemental
Composition
C12H435C140
C12H435C1337C10
13C12H435C140
13C12H435C1337C10
C12H435C1402
C12H435C1337C102
13C12H435C1402
13C12H435C1337C102
C12H435C1537C10
£9^13
C12H335C1437C10
C12H335C1337C120
13C12H335C1437C10
Ci2H3 C 1 3 C 1 20
C12H335C1437C102
C12H335C1337C1202
13C12H335C1437C102
13C12H335C1337C1202
C12H335C1637C10
C9F13
C12H235C1537C10
C12H235C1437C120
13C12H235C160
13C12H235C1537C10
C12H235C1537C102
C12H235C1437C1202
13C12H235C1537C102
13C12H235C1437C1202
C12H235C1637C120
C9F17
Analyte
TCDF
TCDF
TCDF (S)
TCDF (S)
TCDD
TCDD
TCDD (S)
TCDD (S)
HxCDPE
PFK
PeCDF
PeCDF
PeCDF (S)
PeCDF (S)
PeCDD
PeCDD
PeCDD (S)
PeCDD (S)
HpCDPE
PFK
HxCDF
HxCDF
HxCDF (S)
HxCDF (S)
HxCDD
HxCDD
HxCDD (S)
HxCDD (S)
OCDPE
PFK
                      8290 -  60
    Revision 0
September 1994

-------


Descriptor Accurate'8' Ion
Mass ID
4 407.7818 M+2
409.7788 M+4
417.8250 M
419.8220 M+2
423.7767 M+2
425.7737 M+4
435.8169 M+2
437.8140 M+4
479.7165 M+4
[430.9728] LOCK
5 441.7428 M+2
443.7399 M+4
457.7377 M+2
459.7348 M+4
469.7780 M+2
471.7750 M+4
513.6775 M+4
[442.9728] LOCK
Table 6.
Continued
Elemental
Composition
C12H35C1637C10
C12H35C1537C120
13C12H35C170
13C12H35C1637C10
C12H35C1637C102
C12H35C1537C1202
13C12H35C1637C102
13C12H35C1537C1202
C12H35C1737C120
CgF-i?
C1235C1737C10
C1235C1637C120
C1236C1737C102
C1235C1637C1202
13C1235C1737C102
13f 35f1 37f-i «
^12 ^'6 ^'2U2
r 35n 37fi o
^12 UI8 U ' 2U
^10^17


Analyte

HpCDF
HpCDF
HpCDF (S)
HpCDF
HpCDD
HpCDD
HpCDD (S)
HpCDD (S)
NCDPE
PFK
OCDF
OCDF
OCDD
OCDD
OCDD (S)
OCDD (S)
DCDPE
PFK
(8) The following nuclidic masses were used:
H = 1.007825
C = 12.000000
13C = 13.003355
F = 18.9984
0 = 15.994915
35C1 = 34.968853
37C1 = 36.965903





S = internal/recovery standard
                                  8290  - 61
    Revision 0
September 1994

-------
                                   Table 7.

             PCDD  and  PCDF  Congeners  Present  in  the  GC  Performance
                Evaluation Solution  and  Used for  Defining  the
                  Homologous GC Retention Time Windows on a
                               60 m DB-5 Column
No. of
Chlorine
Atoms
4<°>
5
6
7
8
PCDD Positional
First
Eluter
1,3,6,8
1,2,4,6,8/
1,2,4,7,9
1,2,4,6,7,9/
1,2,4,6,8,9
1,2,3,4,6,7,9

Isomer
Last
Eluter
1,2,8,9
1,2,3,8,9
1,2,3,4,6,
1,2,3,4,6,
1,2,3,4,6,
PCDF Positional
First
Eluter
1,3,6,8
1,3,4,6,8
7 1,2,3,4,6,8
7,8 1,2,3,4,6,7,8
7,8,9
Isomer
Last
Eluter
1,2,8,9
1,2,3,8,9
1,2,3,4,8,9
1,2,3,4,7,8,9
1,2,3,4,6,7,8,9
In addition  to  these  two TCDD isomers,  the  1,2,3,4-,  1,2,3,7-, 1,2,3,8-, 2,3,7,8-,
13C12-2,3,7,8-,  and  1,2,3,9-TCDD  isomers must  also  be  present  as  a  check of column
resolution.
                                   8290 - 62                         Revision  0
                                                                 September  1994

-------
                             Table  8.

     Theoretical  Ion Abundance Ratios and Their Control Limits
                        for PCDDs and PCDFs
Number of
Chlorine Ion
Atoms Type
4
5
6
6w
ylb)
7
8
M/M+2
M+2/M+4
M+2/M+4
M/M+2
M/M+2
M+2/M+4
M+2/M+4
Theoretical
Ratio
0.77
1.55
1.24
0.51
0.44
1.04
0.89
Control
lower
0.65
1.32
1.05
0.43
0.37
0.88
0.76
Limits
upper
0.89
1.78
1.43
0.59
0.51
1.20
1.02
M       Used  only for 13C-HxCDF  (IS).

(bl       Used  only for 13C-HpCDF  (IS).
                             8290 - 63
    Revision 0
September 1994

-------
                     Table 9.

Relative Response Factor [RF (number)] Attributions
Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30

2,3,7
2,3,7
1,2,3
1,2,3
2,3,4
1,2,3
1,2,3
1,2,3
1 ,?,3
1,2,3
1,2,3
2,3,4
1,2,3
1,2,3
1,2,3
OCDD
OCDF
13C12-
13p
13C12-
13C12-
13C12-
13C
13C12-
13C12-
13r -
L12-
Total
Total
Total
Total
Specific Congener Name
,8-TCDD (and total TCDDs)
,8-TCDF (and total TCDFs)
,7,8-PeCDD (and total PeCDDs)
,7,8-PeCDF
,7,8-PeCDF
,4,7,8-HxCDD
,6,7,8-HxCDD
,7,8,9-HxCDD
4,7,8-HxCDF
,6,7,8-HxCDF
,7,8,9-HxCDF
,6,7,8-HxCDF
,4,6,7,8-HpCDD (and total HpCDDs)
,4,6,7,8-HpCDF
,4,7,8,9-HpCDF


2,3,7,8-TCDD
2, 3, 7, 8-TCDF
1,2,3,7,8-PeCDD
1,2, 3, 7,8-PeCDF
1,2,3,6,7,8-HxCDD
1,2,3,4,7,8-HxCDF
1,2,3,4,6,7,8-HpCDD
1,2,3,4,6,7,8-HpCDF
OCDD
PeCDFs
HxCDFs
HxCDDs
HpCDFs
                     8290 - 64                         Revision 0
                                                   September 1994

-------
                                Table 10.

         2,3,7,8-TCDD  Toxicity  Equivalency  Factors  (TEFs)  for  the
             Polychlorinated  Dibenzodioxins  and Dibenzofurans
Number           Compound(s)                          TEFa
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
2,3,7,8-TCDD
1,2,3,7,8-PeCDD
1,2,3,6,7,8-HxCDD
1,2,3,7,8,9-HxCDD
1,2,3,4,7,8-HxCDD
1,2,3,4,6,7,8-HpCDD
1,2,3,4,6,7,8,9-OCDD
2,3,7,8-TCDF
1,2,3,7,8-PeCDF
2,3,4,7,8-PeCDF
1,2,3,6,7,8-HxCDF
1,2,3,7,8,9-HxCDF
1,2,3,4,7,8-HxCDF
2,3,4,6,7,8-HxCDF
1,2,3,4,6,7,8-HpCDF
1,2,3,4,7,8,9-HpCDF
1,2,3,4,6,7,8,9-OCDF
1.00
0.50
0.10
0.10
0.10
0.01
0.001
0.1
0.05
0.5
0.1
0.1
0.1
0.1
0.01
0.01
0.001
Taken from "Interim Procedures for Estimating Risks Associated with  Exposures
to Mixtures of Chlorinated Dibenzo-p-Dioxin and -Dibenzofurans (CDDs and CDFs)
and 1989 Update",  (EPA/625/3-89/016, March 1989).
                                8290 -  65                        Revision 0
                                                             September 1994

-------
                                   Table 11.

            Analyte Relative Retention  Time Reference  Attributions
                                                              (a)
            Analyte                     Analyte  RRT  Reference
            1,2,3,4,7,8-HxCDD            13C12-l,2,3,6,7,8-HxCDD

            1,2,3,6,7,8-HxCDF            13C12-l,2,3,4,7,8-HxCDF

            1,2,3,7,8,9-HxCDF            13C12-l,2,3,4,7,8-HxCDF

            2,3,4,6,7,8-HxCDF             13C12-l,2,3,4,7,8-HxCDF
(al  The retention time  of 2,3,4,7,8-PeCDF on the DB-5 column is measured relative
      13C
   to 13C12-l,2,3,4,6,7,8-HpCDF.
to 13C12-l,2,3,7,8-PeCDF and the retention time of 1,2,3,4,7,8,9-HpCDF relative
                                   8290 - 66                         Revision 0
                                                                 September 1994

-------
  a
  o >•
  Q- ae
  is
  is
  U_ LU
  O O.
  UJ 00
  CD 00
  QZ
  UJ O
  Q£0
  O (/)
O —I t5 — •
O> O i— i OO
CsJ O. X Z
00   -v^QC
O
X
    Q- O
  (O CS X
  Q O — *
  Qt-
  o <
  a. £
   .  .O
                                                                               OJ E
                                                                               01 V
                                                                                 4->
                                                                                 Q.
                                                                                 0)
                                                                                 tS>
t
                                                            to
                                                                               CM
                                                                               00
  O
  O
  a.
    >-
    03
                                                     * $
                                                     ll

-------
O
a\ -o
CNJ cu
00 3

Q -r-
O -f->
3C C
                                        -
                            [I

                              I i «B W
                              i--2-^
                                                                                          O TJ-
                                                                                            01
                                                                                          C Ol
                                                                                          O -i
                                                                                          > .0
                                                                                          D £
                                                                                         Q£ O)
                                                                                           +J
                                                                                            Q.
                                                                                            (1)
                                                             a? s S " e c* t ra   c S —
                                                             ?lf illf ? .lis
00
O
01
CM

-------
                                                                                                     I
                                                                                    <0
                                                                                                                                          O ^f
                                                                                                                                             0>
                                                                                                                                          c o>
                                                                                                                                          O •— i
                                                                                                                                          •^
                                                                                                                                          «/l  S-
                                                                                                                                          •r-  0)
                                                                                                                                          > .a
                                                                                                                                          0)  E
                                                                                                                                              a.
                                                                                                                                              0)
CSJ  O>
00  3

O ••-
I—  O
LU  U
 I

O

CM
00

-------
O
O> T3
CM  0)
00  3

Q -r-
O -t->
rc  c
i—  o
LU  O
                                1 -i                 ^
                                                     CO
                         7-  CM  CO
                                                                                          en
                                                                                          en
                                                                                        •f- 0)
                                                                                        > -O
                                                                                        cu E
                                                                                        cr CD
                                                                                          4->
                                                                                          ex
                                                                                          O)
                                                                                          oo
0J
       en
       C\J
       oo

-------
CNJ  0)
00  3
    C
Q •!-
O +J
o:  c
t—  o
UJ  U
         ©1
                                                                     00 m
                                                                     r-'P
                                                                       1
                                                                     i
                                                                       .
                                                                     (0 3
                                                                     o a
                                                                       *
                                                                                                                            o *»•
                                                                                                                               en
                                                                                                                             c o\
                                                                                                                             o i-<

                                                                                                                             M  i-
                                                                                                                            ••-  Q)
                                                                                                                             >.O
                                                                                                                             
-------
4.3  DETERMINATION OF ORGANIC ANALYTES

     4.3.3  HIGH PERFORMANCE LIQUID CHROMATOGRAPHIC  METHODS

         The following methods are included  in  this  section:
         Method 8310:
         Method 8315:

               Appendix A:

         Method 8316:

         Method 8318:

         Method 8321:



         Method 8330:

         Method 8331:
Polynuclear Aromatic Hydrocarbons
Determination  of  Carbonyl   Compounds  by  High
Performance Liquid Chromatography (HPLC)
      Recrystallization    of    2,4-
      Dinitrophenylhydrazine (DNPH)
Acrylamide, Acrylonitrile and  Acrolein by  High
Performance Liquid Chromatography (HPLC)
N-Methylcarbamates  by  High  Performance  Liquid
Chromatography (HPLC)
Solvent  Extractable Non-Volatile  Compounds  by
High    Performance     Liquid
Chromatography/Thermospray/Mass   Spectrometry
(HPLC/TSP/MS)  or Ultraviolet (UV)  Detection
Nitroaromatics and Nitramines by High Performance
Liquid Chromatography  (HPLC)
Tetrazene  by  Reverse   Phase  High   Performance
Liquid Chromatography  (HPLC)
                                 FOUR  -  12
                                      Revision  2
                                  September  1994

-------
                                 METHOD  8310

                      POLYNUCLEAR AROMATIC  HYDROCARBONS
1.0  SCOPE AND APPLICATION

     1.1  Method 8310 is used to  determine the concentration of certain poly-
nuclear aromatic hydrocarbons (PAH)  in ground water and wastes.   Specifically,
Method 8310 is used to detect the following substances:

          Acenaphthene                     Chrysene
          Acenaphthylene                   Di benzo(a,h)anthracene
          Anthracene                       Fluoranthene
          Benzo(a)anthracene               Fluorene
          Benzo(a)pyrene                   Indeno(1,2,3-cd)pyrene
          Benzo(b)fluoranthene             Naphthalene
          Benzo(ghi)perylene               Phenanthrene
          Benzo(k)fluoranthene             Pyrene

     1.2  Use of Method 8310  presupposes  a  high  expectation of finding the
specific compounds of interest.  If  the  user is attempting to screen samples
for any or all  of  the  compounds  listed  above, he must develop independent
protocols for the verification of identity.

     1.3  The method detection limits for  each  compound in reagent water are
listed  in Table 1.  Table  2  lists the practical quantitation limit (PQL) for
other matrices.  The sensitivity of  this  method usually depends on the level
of  interferences  rather  than  instrumental  limitations.    The  limits  of
detection listed in Table 1  for the liquid chromatographic approach represent
sensitivities that can be  achieved  in  the  absence  of interferences.  When
interferences are present, the level of sensitivity will be lower.

     1.4  This method  is  recommended  for  use  only  by experienced residue
analysts or under the close  supervision of such qualified persons.


2.0  SUMMARY OF METHOD

     2.1  Method 8310 provides high  performance liquid chromatographic  (HPLC)
conditions for the detection  of  ppb  levels  of certain polynuclear aromatic
hydrocarbons.  Prior  to  use  of  this  method, appropriate sample extraction
techniques must be used.  A  5-  to  25-uL  aliquot of the extract is injected
into an HPLC, and compounds  in  the  effluent are detected by ultraviolet (UV)
and fluorescence detectors.

     2.2  If  interferences  prevent  proper  detection  of  the  analytes  of
interest, the method may  also  be  performed  on extracts that have undergone
cleanup using silica gel column cleanup  (Method 3630).
                                  8310 -  1
                                                         Revision
                                                         Date  September 1986

-------
TABLE 1.  HIGH PERFORMANCE LIQUID CHROMATOGRAPHY OF PAHsa
Compound
Naphthalene
Acenaphthylene
Acenaphthene
Fluorene
Phenanthrene
Anthracene
Fluoranthrene
Pyrene
Benzo (a) anthracene
Chrysene
Benzo (b) f 1 uoranthene
Benzo (k) f 1 uoranthene
Benzo(a)pyrene
D1 benzo (a , h) anthracene
Benzo (ghi)perylene
Indeno(l,2,3-cd)pyrene
Retention
time (mln)
16.6
18.5
20.5
21.2
22.1
23.4
24.5
25.4
28.5
29.3
31.6
32.9
33.9
35.7
36.3
37.4
Col umn
capacity
factor (k1)
12.2
13.7
15.2
15.8
16.6
17.6
18.5
19.1
21.6
22.2
24.0
25.1
25.9
27.4
27.8
28.7
Method Detection
limit (ug/L)
UV Fluorescence
1.8
2.3
1.8
0.21
0.64
0.66
0.21
0.27
0.013
0.15
0.018
0.017
0.023
0.030
0.076
0.043
      a  HPLC  conditions:   Reverse   phase   HC-ODS S11-X,  5 micron particle  size,
 1n  a  250-mm  x 2.6-mrn I.D.  stainless  steel  column.   Isocratlc elutlon  for  5 mln
 using acetom'trlle/water  (4:6)(v/v),  then   linear  gradient  elutlon  to 100%
 acetonltrlle over 25 mln at  0.5   mL/min   flow rate.   If  columns  having  other
 Internal  diameters are used,  the   flow rate   should  be adjusted to maintain  a
 linear  velocity of 2 mm/sec.
 TABLE 2.   DETERMINATION OF PRACTICAL QUANTITATION LIMITS (PQL)  FOR VARIOUS
           MATRICES3
     Matrix
  Factor*3
 Ground water
 Low-level  soil  by sonication with GPC cleanup
 High-level  soil and sludges by sonlcatlon
 Non-water mlsdble waste
     10
    670
 10,000
100,000
      aSample PQLs are highly  matrix-dependent.     The  PQLs listed herein are
      provided for guidance and may not always be achievable.
           = [Method Detection Limit (Table 1)   X [Factor (Table 2)].  For non-
      aqueous samples, the factor is on a wet-weight basis.
                                   8310 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
3.0  INTERFERENCES

     3.1  Solvents, reagents, glassware, and  other sample processing hardware
may yield discrete artifacts and/or elevated baselines, causing misinterpreta-
tion of the chromatograms.  All of  these materials must be demonstrated to be
free from interferences,  under  the  conditions  of  the analysis, by running
method blanks.  Specific selection of reagents and purification of solvents by
distillation in all-glass systems may be required.

     3.2  Interferences coextracted from  the  samples  will vary considerably
from source to source.   Although  a  general cleanup technique is provided as
part  of  this  method,  individual  samples  may  require  additional cleanup
approaches to achieve the sensitivities stated in Table 1.

     3.3  The  chromatographic  conditions   described   allow  for  a  unique
resolution of the specific PAH  compounds  covered  by this method.  Other PAH
compounds, in addition to matrix artifacts, may interfere.


4.0  APPARATUS AND MATERIALS

     4.1  Kuderna-Danish  (K-D) apparatus:

          4.1.1  Concentrator  tube:  10-ml, graduated  (Kontes K-570050-1025 or
     equivalent).  Ground-glass  stopper  is  used  to  prevent evaporation of
     extracts.

          4.1.2  Evaporation   flask:      500-mL   (Kontes   K-570001-500  or
     equivalent).  Attach to concentrator tube with springs.

          4.1.3  Snyder column:     Three-ball  macro   (Kontes K-503000-0121 or
     equivalent).

          4.1.4  Snyder  column:    Two-ball  micro   (Kontes  K-569001-0219 or
     equivalent).

     4.2  Boiling  chips;  Solvent  extracted,  approximately  10/40  mesh  (silicon
 carbide or equivalent).

     4.3  Water  bath;    Heated,   with concentric   ring   cover,   capable  of
 temperature  control  (+5'C).  The bath  should  be  used  in  a  hood.

     4.4  Syringe;   5-mL.

     4.5  High pressure syringes.

     4.6  HPLC apparatus;

          4.6.1  Gradient pumping  system:  Constant flow.

          4.6.2  Reverse  phase column:   HC-ODS   Sil-X,  5-micron  particle  size
     diameter, in  a  250-mm  x 2.6-mm  I.D.  stainless steel  column  (Perkin Elmer
     No.  089-0716  or equivalent).

                                   8310  - 3
                                                         Revision      0
                                                          Date  September  1986

-------
          4.6.3  Detectors:   Fluorescence  and/or UV detectors may be used.

               4.6.3.1   Fluorescence  detector:   For   excitation at 280-nm and
          emission greater than  389-nm cutoff   (Corning  3-75 or equivalent).
          Fluorometers  should have dispersive  optics  for  excitation  and can
          utilize either filter or dispersive optics  at  the  emission detector.

               4.6.3.2   UV detector:     254-nm,  coupled  to the fluorescence
          detector.

          4.6.4  Strip-chart recorder:   compatible  with  detectors.     A data
     system for measuring peak areas  and retention  times is  recommended.

     4.7  Volumetric flasks;  10-, 50-, and 100-mL.


5.0  REAGENTS

     5.1  Reagent water:   Reagent water   is  defined  as  water   1n which  an
interferent is not observed at the method detection  limit of the compounds  of
interest.

     5.2  Acetonitrile;  HPLC quality,  distilled  in glass.

     5.3  Stock standard solutions;

          5.3.1  Prepare stock standard solutions  at  a concentration  of 1.00
     ug/uL by dissolving  0.0100  g  of  assayed   reference  material  in aceto-
     nitrile and diluting  to  volume  in  a  10-mL  volumetric flask.   Larger
     volumes can be used at  the  convenience  of  the analyst.  When compound
     purity is assayed to be 96%  or  greater,   the weight can  be used without
     correction  to  calculate  the   concentration  of  the  stock  standard.
     Commercially  prepared  stock  standards can be used at any concentration if
     they are  certified by  the manufacturer or by an  independent source.

          5.3.2  Transfer   the  stock  standard  solutions  into Teflon-sealed
     screw-cap bottles.  Store at 4*C  and protect from light.  Stock standards
     should be checked   frequently   for   signs  of degradation or evaporation,
     especially  just prior  to  preparing calibration  standards from them.

          5.3.3   Stock  standard solutions must be  replaced after one year, or
     sooner  if comparison with check standards indicates  a  problem.

     5.4 Calibration  standards;   Calibration  standards  at a minimum of five
 concentration   levels   should   be prepared  through dilution  of  the  stock
 standards with acetonitrile.   One of  the concentration  levels should be at a
 concentration  near,  but  above,   the  method  detection   limit.  The remaining
 concentration  levels should correspond to the expected  range of concentrations
 found  in real  samples  or should define the  working range  of the HPLC.  Cali-
 bration standards must be replaced  after six months,  or sooner if comparison
 with check  standards indicates a  problem.
                                   8310 - 4
                                                          Revision      0
                                                          Date  September  1986

-------
     5.5  Internal  standards (1f Internal   standard  calibration 1s used);   To
use this approach,  the analyst must select one or more internal  standards that
are similar in analytical behavior to  the compounds of interest.  The analyst
must further demonstrate that the measurement  of the internal  standard is not
affected by method or matrix interferences.   Because of these limitations, no
internal standard can be suggested that is applicable to all samples.

          5.5.1  Prepare  calibration   standards   at   a   minimum  of  five
     concentration levels for each analyte as described in Paragraph 5.4.

          5.5.2  To each calibration standard, add  a known constant amount of
     one or more internal standards, and dilute to volume with acetonitrlle.

          5.5.3  Analyze each calibration standard according to Section 7.0.

     5.6  Surrogate standards;  The analyst  should monitor the performance of
the  extraction,cleanup(Tf  necessary),  and  analytical  system  and  the
effectiveness of the method in dealing with each sample matrix by spiking each
sample, standard, and reagent water  blank  with  one or two surrogates (e.g.,
decafluorobiphenyl or other PAHs  not  expected  to  be present in the sample)
recommended to encompass the  range  of  the  temperature program used in this
method.  Deuterated analogs of analytes  should  not be used as surrogates for
HPLC analysis due to coelutlon problems.


6.0  SAMPLE COLLECTION,  PRESERVATION, AND HANDLING

     6.1  See the introductory  material  to  this  chapter, Organic Analytes,
Section 4.1.  Extracts must be stored under refrigeration and must be analyzed
within 40 days of extraction.


7.0  PROCEDURE

     7.1  Extraction;

          7.1.1  Refer to Chapter Two for guidance on choosing the appropriate
     extraction procedure.   In  general,  water  samples  are  extracted at a
     neutral  pH with methylene  chloride,  using  either  Method 3510 or 3520.
     Solid  samples are extracted using either Method 3540 or 3550.   To achieve
     maximum  sensitivity with this method, the extract must be concentrated to
     1 mL.

          7.1.2  Prior   to  HPLC  analysis,  the  extraction  solvent  must be
     exchanged to acetonitrile.    The  exchange  1s  performed during the K-D
     procedures  listed  in all  of  the  extraction  methods.   The exchange 1s
     performed as follows.

               7.1.2.1   Following K-D of the methylene chloride extract to
           1 mL using the macro-Snyder column,  allow the apparatus to cool and
          drain  for at  least  10 min.
                                   8310 -  5
                                                          Revision      0
                                                          Date  September 1986

-------
         7.1.2.2   Increase the temperature of the  hot water bath to 95-
     100'C.    Momentarily  remove  the   Snyder  column,  add  4  ml  of
     acetonltrlle,  a new boiling chip, and attach a two-ball mlcro-Snyder
     column.  Concentrate  the  extract  using  1  ml  of acetonltrlle to
     prewet the  Snyder  column.  Place the K-D apparatus on the water bath
     so  that  the  concentrator  tube  1s  partially  Immersed  1n the hot
     water.   Adjust the vertical position  of the apparatus and the water
     temperature,  as required,  to  complete  concentration 1n 15-20 m1n.
     At  the proper rate of   distillation  the  balls  of the column will
     actively chatter,  but   the  chambers  will  not  flood.    When the
     apparent volume of liquid  reaches  0.5  ml, remove the K-D apparatus
     and allow  1t  to drain and  cool for at least 10 m1n.

          7.1.2.3   When the apparatus   is  cool,  remove the micro-Snyder
     column  and rinse  its  lower  joint   Into   the concentrator tube with
     about 0.2  ml  of   acetonltrlle.     A  5-mL  syringe is recommended for
     this operation.   Adjust  the  extract  volume  to  1.0 ml.  Stopper the
     concentrator   tube  and   store    refrigerated  at  4*C,  if   further
     processing will  not  be performed   immediately.    If the extract will
     be stored  longer  than   two  days,   it  should   be  transferred to  a
     Teflon-sealed screw-cap  vial.   Proceed with HPLC analysis  1f  further
     cleanup is not required.

7.2  HPLC conditions  (Recommended);

     7.2.1  Using  the column   described  in  Paragraph  4.6.2:   Isocratlc
elution  for  5  m1n   using   aceton1tr1le/water  (4:6)(v/v),  then linear
gradient elution to 100%   acetonitrile  over  25  min at 0.5 mL/min flow
rate.  If columns  having  other Internal  diameters are used,  the flow rate
should be adjusted to maintain a  linear velocity of 2 mm/sec.

7.3  Calibration:

     7.3.1  Refer to Method  8000  for  proper  calibration procedures.   The
procedure of internal  or external standard   calibration  may be  used.   Use
Table 1 and especially Table 2 for guidance on  selecting the lowest point
on the calibration curve.

     7.3.2  Assemble the necessary HPLC apparatus  and establish operating
parameters equivalent to those indicated  in  Section  7.2.1.  By injecting
calibration standards, establish the  sensitivity  limit of the detectors
and  the linear  range of the analytical systems  for each  compound.

     7.3.3  Before  using  any  cleanup  procedure,   the  analyst  should
process a  series  of  calibration  standards  through  the  procedure to
confirm elution patterns  and  the  absence  of  Interferences  from the
reagents.

7.4  HPLC analysis;

     7.4.1   Table  1  summarizes  the  estimate  retention  times  of PAHs
determinable by this method.   Figure  1  1s an example of the separation
achievable using  the conditions given in Paragraph 7.2.1.

                             8310 - 6
                                                    Revision      0
                                                    Date  September 1986

-------
Column: HC-ODSSIL-X
Mobile Phase: 40% to 100% Acetonitrile in Water
Dectector: Fluorescence
                                                       40
                RETENTION TIME (MINUTES)
        Figure 1.  Liquid chromatogram of poly nuclear aromatics.
                        8310 - 7
                                                 Revision       p
                                                 Date  September 1986

-------
         7.4.2  If Internal standard calibration Is to be performed, add
    10 uL of Internal standard to the sample prior to Injection.  Inject
    2-5 uL of  the  sample  extract  with  a  high-pressure syringe or sample
    injection loop.  Record the  volume  Injected  to the nearest 0.1 uL, and
    the resulting peak size, in  area  units or peak heights.  Re-equilibrate
    the HPLC column at the  initial  gradient  conditions for at least 10 min
    between injections.

         7.4.3  Using either the  internal  or external calibration procedure
     (Method 8000), determine the identity and quantity of each component peak
     in the sample chromatogram  which  corresponds  to the compounds used for
    calibration purposes.  See  Section  7.8  of  Method 8000 for calculation
     equations.

         7.4.4   If the peak  area   exceeds  the  linear  range of the system,
     dilute the extract and  .^analyze.

         7.4.5   If the peak  area measurement  is prevented by the presence of
     interferences, further  cleanup  is  required.

     7.5  Cleanup:

         7.5.1   Cleanup  of  the  acetonitrile   extract  takes place using Method
     3630 (Silica Gel   Cleanup).     Specific   instructions  for  cleanup of the
     extract  for PAHs is  given  in  Section 7.1  of Method 3630.

          7.5.2  Following  cleanup,  analyze    the    samples  using  HPLC  as
     described in Section 7.4.
8.0  QUALITY CONTROL

     8.1  Refer  to  Chapter  One  for  specific  quality  control procedures.
Quality control to validate sample extraction is covered in Method 3500 and in
the extraction method used.  If  extract  cleanup was performed,  follow the QC
in Method 3600 and in the specific cleanup method.

     8.2  Mandatory quality control to  validate  the HPLC system operation is
found 1n Method 8000, Section 8.6.

          8.2.1  The quality control  check  sample  concentrate  (Method 8000,
     Section 8.6) should contain each  analyte at the following concentrations
     in acetonitrile:    naphthalene,  100  ug/mL;  acenaphthylene, 100 ug/mL;
     acenaphthene, 100 ug/mL;  fluorene,  100  ug/mL; phenanthrene, 100 ug/mL;
     anthracene, 100 ug/mL; benzo(k)fluoranthene,  5  ug/mL; and any other PAH
     at 10 ug/mL.

          8.2.2  Table 3 indicates the  calibration and QC acceptance criteria
     for this  method.    Table  4  gives  method  accuracy  and  precision as
     functions of concentration for the analytes of interest.  The contents of
     both Tables should be used to  evaluate a laboratory's ability to perform
     and generate acceptable data by this method.


                                  8310 - 8
                                                         Revision      0
                                                         Date  September 1986

-------
     8.3  Calculate surrogate standard  recovery  on  all samples, blanks, and
spikes.  Determine if  the  recovery  is  within limits (limits established by
performing QC procedures outlined in Method 8000, Section 8.10).

          8.3.1  If recovery is  not  within  limits, the following procedures
     are required.

               •  Check to  be  sure  there  are  no  errors  in calculations,
                  surrogate solutions  and  internal  standards.   Also, check
                  instrument performance.

               •  Recalculate the data and/or reanalyze  the extract if any of
                  the above checks reveal a problem.

               •  Reextract and reanalyze the sample  if none of the above are
                  a problem or flag the data as  "estimated concentration."


9.0  METHOD  PERFORMANCE

     9.1  The method  was  tested  by  16  laboratories  using  reagent water,
drinking water,  surface water, and  three industrial wastewaters  spiked at six
concentrations over the range  0.1  to  425  ug/L.   Single operator precision,
overall precision, and method accuracy  were  found  to  be directly related to
the  concentration of the   analyte  and  essentially  independent  of the sample
matrix.  Linear  equations  to  describe  these   relationships  are presented in
Table  4.

     9.2  This method has  been  tested  for  linearity   of spike  recovery from
reagent  water   and  has   been   demonstrated    to  be  applicable  over the
concentration  range from 8 x MDL  to  800  x MDL with  the following exception:
benzo(ghi)perylene recovery at 80  x  and  800   x MDL   were  low  (35%  and 45%,
respectively).

     9.3  The  accuracy  and precision  obtained will  be  determined  by the sample
matrix,  sample-preparation technique, and calibration  procedures  used.


10.0  REFERENCES

1.   "Development and Application of Test  Procedures  for  Specific  Organic  Toxic
Substances  in  Wastewaters, Category 9 -  PAHs,"   Report for EPA Contract 68-03-
2624 (in preparation).

2.   Sauter,  A.D.,  L.D.  Betowski, T.R. Smith, V.A. Strickler,  R.6. Beimer, B.N.
Colby, and   J.E. Wilkinson,   "Fused  Silica  Capillary  Column  GC/MS for the
Analysis of Priority Pollutants," Journal of HRC&CC  4, 366-384, 1981.

3.   "Determination of   Polynuclear  Aromatic  Hydrocarbons   in  Industrial and
Municipal   Wastewaters,"    EPA-600/4-82-025,    U.S.   Environmental Protection
Agency,  Environmental   Monitoring  and  Support  Laboratory,   Cincinnati, Ohio
45268, September 1982.


                                  8310 -  9
                                                         Revision      0
                                                         Date  September  1986

-------
4.  Burke, J.A.  "Gas  Chromatography  for  Pesticide  Residue  Analysis; Some
Practical  Aspects,"  Journal  of   the  Association  of  Official  Analytical
Chemists, 48, 1037, 1965.

5.  "EPA  Method  Validation  Study   20,  Method  610  (Polynuclear  Aromatic
Hydrocarbons)," Report for EPA Contract 68-03-2624 (in preparation).

6.  U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the
Analysis of Pollutants Under the Clean Water Act; Final Rule and  Interim Final
Rule and Proposed Rule," October 26, 1984.

7.  Provost, L.P. and R.S.  Elder,  "Interpretation of Percent Recovery Data,"
American Laboratory, Ijj, pp. 58-63, 1983.
                                   8310 - 10
                                                          Revision
                                                          Date  September 1986

-------
TABLE 3.  QC ACCEPTANCE CRITERIA*
Parameter
Acenaphthene
Acenaphthylene
Anthracene
Benzo (a) anthracene
Benzo (a) pyrene
Benzo (b) f 1 uoranthene
Benzo (ghi)perylene
Benzo (k) f 1 uoranthene
Chrysene
01 benzo (a , h) anthracene
Fl uoranthene
Fluorene
Indeno (1 , 2 , 3-cd) pyrene
Naphthalene
Phenanthrene
Pyrene
Test
cone.
(ug/L)
100
100
100
10
10
10
10
5
10
10
10
100
10
100
100
10
Limit
for s
(ug/L)
40.3
45.1
28.7
4.0
4.0
3.1
2.3
2.5
4.2
2.0
3.0
43.0
3.0
40.7
37.7
3.4
Range
for X
(ug/L)
D-105.7
22.1-112.1
11.2-112.3
3.1-11.6
0.2-11.0
1.8-13.8
D-10.7
D-7.0
D-17.5
0.3-10.0
2.7-11.1
D-119
1.2-10.0
21.5-100.0
8.4-133.7
1.4-12.1
Range
P. Ps
(%)
D-124
D-139
D-126
12-135
D-128
6-150
D-116
D-159
D-199
D-110
14-123
D-142
D-116
D-122
D-155
D-140
     s = Standard deviation of four recovery measurements, in ug/L.
     7 = Average recovery for four recovery measurements, in ug/L.
     p, ps = Percent recovery measured.
     D = Detected; result must be greater than zero.
     aCriteria from 40 CFR Part  136 for  Method 610.  These criteria are based
directly upon the method performance  data  in  Table 3.  Where necessary, the
limits for recovery have been broadened  to assure applicability of the limits
to  concentrations below those used to develop Table 3.
                                   8310 - 11
                                                          Revision       0
                                                          Date  September  1986

-------
TABLE 4.  METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATIONa
Parameter
Acenaphthene
Acenaphthylene
Anthracene
Benzo (a) anthracene
Benzo(a)pyrene
Benzo (b) f 1 uoranthene
Benzo (ghi)perylene
Benzo (k) f 1 uoranthene
Chrysene
Dlbenzo (a, h) anthracene
Fl uoranthene
Fluorene
Indeno (1 , 2 , 3-cd)pyrene
Naphthalene
Phenanthrene
Pyrene
Accuracy, as
recovery, x1
(ug/L)
0.52C+0.54
0.69C-1.89
0.63C-1.26
0.73C+0.05
0.56C+0.01
0.78C+0.01
0.44C+0.30
0.59C+0.00
0.77C-0.18
0.41C-0.11
0.68C+0.07
0.56C-0.52
0.54C+0.06
0.57C-0.70
0.72C-0.95
0.69C-0.12
Single analyst
precision, sr'
(ug/L)
0.397+0.76
0.367+0.29
0.237+1.16
0.287+0.04
0.387-0.01
0.217+0.01
0.257+0.04
0.447-0.00
0.327-0.18
0.247+0.02
0.227+0.06
0.447-1.12
0.297+0.02
0.397-0.18
0.297+0.05
0.257+0.14
Overal 1
precision,
S1 (ug/L)
0.537+1.32
0.427+0.52
0.417+0.45
0.347+0.02
0.537-0.01
0.387-0.00
0.587+0.10
0.697+0.10
0.667-0.22
0.457+0.03
0.327+0.03
0.637-0.65
0.427+0.01
0.417+0.74
0.477-0.25
0.427-0.00
      x'   =  Expected   recovery   for  one  or  more  measurements  of  a  sample
            containing a  concentration of C,  in ug/L.

      sr'  =  Expected  single  analyst  standard deviation  of measurements at an
            average concentration  of 7, in  ug/L.

      S1   =  Expected  interlaboratory standard deviation  of measurements at an
            average concentration  found of  7, in  ug/L.

      C    =  True value for the  concentration, 1n  ug/L.

      7    =  Average recovery found for measurements of  samples  containing a
            concentration of C,  in ug/L.
                                   8310 - 12
                                                          Revision      0
                                                          Date  September 1986

-------
                                         METHOD 8310

                              POLYNUCLEAR AROMATIC HYDROCARBONS
 7.1.1
                                                        0
       Choose
    appropriate
    •xtractlon
     procedure
 (see Chapter 2)
7. l.Z
                                                    7.3.3
       Process
    a series  of
    calibration
     standards
       Exchange
       extract-
 ion solvent to
   acetonltrlle
    durlne K-D
    procedures
 7.2
                                                     7.4
       Perform
        HPLC
  analysis  (see
   Method 6000
for calculation
    equations
    Set HPLC
   conditions
 7.3
       Refer to
    Method 8OOO
    for proper
    calibration
    techniques
                          Cleanup using
                           Method 3630
7.3.2
HPLC apparatus:
     establish
     operating
    parameters
    o
                                     8310  - 13
                                                                Revision       o
                                                                Date  September 1986

-------
                                  METHOD  8315

                      DETERMINATION  OF  CARBONYL  COMPOUNDS
               BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)
1.0   SCOPE AND APPLICATION

      1.1   This  method provides  procedures  for  the  determination of  free
carbonyl   compounds    in    various    matrices    by    derivatization    with
2,4-dinitrophenylhydrazine (DNPH).   The  method  utilizes high performance liquid
chromatography (HPLC) with ultraviolet/visible (UV/vis) detection to identify and
quantitate the target analytes  using two different sets of conditions: Option 1
and Option 2.   Option  1 has  been shown to perform well  for  one set of target
analytes for aqueous samples, soil or waste samples, and stack samples collected
by Method 0011.  Option 2 has been  shown to work well for another set of target
analytes in indoor air  samples collected by Method 0100.   The two sets of target
analytes overlap for some compounds.  Refer to the Analysis Option listed in the
following table to determine which analytes may  be analyzed by which  option.  The
following compounds may be determined by this method:
      Compound Name                       CAS No.a          Analysis Optionb
Acetaldehyde
Acetone
Acrolein
Benzaldehyde
Butanal (butyraldehyde)
Crotonaldehyde
Cyclohexanone
Decanal
2,5-Dimethylbenzaldehyde
Formaldehyde
Heptanal
Hexanal (hexaldehyde)
Isovaleraldehyde
Nonanal
Octanal
Pentanal (valeraldehyde)
Propanal (propionaldehyde)
m-Tolualdehyde
o-Tolualdehyde
p-Tolualdehyde
a Chemical Abstract Services
" Tk-io 1 -i o +• rv-P + •» v»n r\4- •!» m •* 1
75-07-0
67-64-1
107-02-8
100-52-7
123-72-8
123-73-9
108-94-1
112-31-2
5779-94-2
50-00-0
111-71-7
66-25-1
590-86-3
124-19-6
124-13-0
110-62-3
123-38-6
620-23-5
529-20-4
104-87-0
Registry Number.
\/4- n o /->/MTt-^-ir\r* •% rv n
1,2
2
2
2
1,2
1,2
1
1
2
1,2
1
1,2
2
1
1
1,2
1,2
2
2
2
klf^\V*l^l"Vr\TM/'t 1 T f» 41 f\ f
            compounds  that  have  been  evaluated  using  modifications of  the
                                   8315 - 1                         Revision 0
                                                                September 1994

-------
            analysis.  Refer to the respective option number when choosing the
            appropriate analysis technique for a particular compound.

      1.2   The Option 1 method detection limits  (MDL) are listed in Tables 1 and
2.  The sensitivity data for sampling  and  analysis using Method 0100  (Option 2)
are given in Table 3.  The MDL for a specific sample may differ from that listed,
depending upon the nature of interferences in the sample matrix and the amount
of sample used in the procedure.

      1.3   The  extraction  procedure  for solid  samples  is  similar  to  that
specified in Method 1311.  Thus, a single sample may  be extracted to measure the
analytes included  in  the  scope of other  appropriate methods.   The  analyst is
allowed the flexibility to select chromatographic conditions appropriate for the
simultaneous measurement of combinations of these analytes.

      1.4   When this  method is used  to analyze unfamiliar  sample matrices,
compound  identification  should  be   supported   by   at  least  one  additional
qualitative technique.  A gas chromatograph/mass spectrometer (GC/MS)  may be used
for the qualitative confirmation of results for the  target analytes, using the
extract produced by this method.

      1.5   This method is  restricted  to  use  by,  or under the supervision of,
analysts experienced in the use of chromatography and in the interpretation of
chromatograms.  Each analyst must demonstrate the ability to generate acceptable
results with this method, using the procedure described in Sec.  7.0.


2.0   SUMMARY OF METHOD

      2.1   Liquid  and Solid Samples (Option 1)

            2.1.1 For  wastes  comprised  of  solids,   or   for  aqueous  wastes
      containing significant  amounts  of solid  material,  the aqueous phase, if
      any, is separated from the solid phase  and  stored  for later analysis.  If
      necessary, the particle size of the solids in  the waste is reduced.  The
      solid phase  is extracted  with an amount  of extraction fluid equal  to 20
      times the weight of the solid phase.  The extraction fluid employed is a
      function of  the  alkalinity  of  the solid phase  of  the waste.   A special
      extractor vessel is  used when testing for volatiles.  Following extraction,
      the  aqueous   extract  is  separated  from the  solid  phase  by   filtration
      employing 0.6 to 0.8  jum glass fiber filter.

            2.1.2  If  compatible  (i.e.,  multiple  phases  will  not  form  on
      combination),  the initial aqueous  phase  of  the  waste  is added to the
      aqueous  extract,   and  these   liquids  are   analyzed   together.     If
      incompatible,  the liquids are  analyzed  separately  and  the  results are
      mathematically combined to yield a volume-weighted  average concentration.

            2.1.3 A measured  volume  of aqueous sample  (approx.  100 mL)  or an
      appropriate  amount of solids extract (approx.  25 g),  is buffered to pH 3
      and derivatized with  2,4-dinitrophenylhydrazine  (DNPH), using  either the
      liquid-solid  or  a liquid-liquid  extraction option.   If the liquid-solid


                                   8315  - 2                          Revision 0
                                                                September 1994

-------
      option  is  used,  the  derivative  is  extracted  using  solid  sorbent
      cartridges, followed by elution with ethanol.   If  the liquid-liquid option
      is  used,  the derivative  is extracted  from  the  sample with  three  (3)
      portions  of  methylene  chloride.   The  methylene chloride  extracts  are
      concentrated using the Kuderna-Danish  (K-D) procedure and  exchanged with
      acetonitrile  prior to HPLC analysis.   Liquid chromatographic conditions
      are described  which  permit  the separation  and  measurement  of various
      carbonyl compounds in the extract by absorbance detection at 360 nm.

            2.1.4 If formaldehyde is the only analyte of interest, the aqueous
      sample or solids extract should be buffered to pH  5.0 to minimize artifact
      formaldehyde formation.

      2.2   Stack Gas Samples Collected by Method 0011  (Option 1) - The entire
sample returned to the laboratory is extracted with methylene chloride and the
methylene chloride extract  is brought up to a known volume.   An aliquot of the
methylene chloride extract  is solvent exchanged and concentrated or diluted as
necessary.   Liquid chromatographic conditions  are described that  permit  the
separation  and  measurement  of  various  carbonyl compounds  in the  extract  by
absorbance detection at 360 nm.

      2.3   Indoor Air Samples by Method 0100 (Option 2) - The  sample cartridges
are returned  to  the  laboratory  and backflushed with acetonitrile  into  a  5 mL
volumetric flask.   The eluate is brought up to volume with more acetonitrile.
Two (2) aliquots of the  acetonitrile  extract  are pipetted  into two (2)  sample
vials having Teflon-lined septa.  Liquid chromatographic conditions are described
that permit the separation and measurement of  the various carbonyl compounds in
the extract by absorbance detection at 360 nm.


3.0   INTERFERENCES

      3.1   Method interferences  may be caused by contaminants  in solvents,
reagents, glassware,  and other sample  processing hardware that lead to discrete
artifacts and/or elevated baselines in the chromatograms.  All  of these materials
must be routinely demonstrated to be free from interferences under  the conditions
of the analysis  by  analyzing laboratory reagent blanks as described in Sec.  8.5.

            3.1.1 Glassware must be scrupulously cleaned.   Clean all glassware
      as soon as possible after use by rinsing with  the  last solvent  used.  This
      should  be  followed by detergent washing  with hot water, and rinses  with
      tap water  and  organic-free reagent water.   It  should  then  be drained,
      dried,  and heated in  a  laboratory oven at  130°C for several hours before
      use.  Solvent rinses  with acetonitrile   may  be  substituted for the  oven
      heating.  After drying and  cooling, glassware should be  stored in a clean
      environment to prevent  any accumulation of dust or other contaminants.

            NOTE: Do not  use acetone  or methanol.   These solvents  react  with
                  DNPH to form interfering compounds.
                                   8315 - 3                         Revision 0
                                                                September 1994

-------
            3.1.2 The use of high purity reagents and solvents helps to minimize
      interference problems.  Purification  of  solvents  by distillation in all
      glass systems may be required.

            3.1.3 Polyethylene gloves must be worn when handling the silica gel
      cartridges to reduce the possibility of contamination.

      3.2   Formaldehyde  contamination  of  the DNPH  reagent  is  a  frequently
encountered problem due to  its widespread  occurrence in  the environment.   The
DNPH reagent in Option 2 must be purified by multiple recrystallizations in UV-
grade acetonitrile.   Recrystallization is  accomplished, at 40-60°C,  by  slow
evaporation of the solvent to maximize crystal  size.   The  purified DNPH crystals
are stored under UV-grade acetonitrile until use.  Impurity levels of carbonyl
compounds in the DNPH are determined  prior  to  the analysis  of  the samples and
should be  less than  25 mg/L.   Refer to Appendix A  for  the recrystallization
procedure.

      3.3   Matrix  interferences   may  be   caused  by  contaminants  that  are
coextracted from  the sample.   The extent  of  matrix interferences  will  vary
considerably from source  to source, depending upon the nature and diversity of
the matrix being sampled.   Although  the HPLC  conditions  described allow for a
resolution  of  the specific compounds  covered by  this  method,  other matrix
components  may  interfere.    If interferences  occur  in subsequent  samples,
modification of the mobile phase or some additional  cleanup may be necessary.

      3.4   In Option 1, acetaldehyde is  generated during the  derivatization step
if ethanol  is present in the sample.  This background  will  impair the measurement
of acetaldehyde at levels below 0.5 ppm (500 ppb).

      3.5   For Option  2, at the  stated  two  column  analysis  conditions,  the
identification  and  quantitation  of  butyraldehyde   may  be  difficult   due  to
coelution with isobutyraldehyde  and methyl  ethyl  ketone.  Precautions should be
taken and adjustment  of  the  analysis conditions should be  done, if necessary, to
avoid potential problems.


4.0   APPARATUS AND MATERIALS

      4.1   High performance liquid chromatograph (modular)

            4.1.1 Pumping system -  Gradient, with constant flow control capable
      of 1.50 mL/min.

            4.1.2 High pressure injection valve with 20 /zL  loop.

            4.1.3 Column  - 250 mm x 4.6 mm ID,  5 /zm particle size, CIS  (Zorbax
      or equivalent).

            4.1.4 Absorbance detector - 360 nm.

            4.1.5 Strip-chart recorder compatible with detector  -  Use of a data
      system for measuring peak areas and retention  times is recommended.


                                   8315 - 4                         Revision 0
                                                                September 1994

-------
      4.1.6 Helium  -  for  degassing  mobile  phase  solvents.  (Options
1 and 2)
      4.1.7 Mobile Phase Reservoirs and Suction Filtration Apparatus  - For
holding and filtering HPLC mobile phase.  Filtering system should be all
glass and Teflon and use a 0.22 /urn polyester membrane filter. (Option 2)
      4.1.8 Syringes - for HPLC injection loop loading, with capacity at
least four times the loop volume.
4.2   Apparatus and Materials for Option 1
      4.2.1 Reaction vessel - 250 ml Florence flask.
      4.2.2 Separatory funnel - 250 ml,  with Teflon stopcock.
      4.2.3 Kuderna-Danish (K-D) apparatus.
            4.2.3.1     Concentrator  tube  -   10  ml  graduated  (Kontes
      K-570050-1025 or  equivalent).   A  ground glass stopper is  used to
      prevent evaporation of extracts.
            4.2.3.2     Evaporation flask - 500 ml  (Kontes K-570001-500 or
      equivalent).  Attach to concentrator tube with springs, clamps, or
      equivalent.
            4.2.3.3     Snyder  column   -  Three   ball  macro   (Kontes
      K-503000-0121 or equivalent).
            4.2.3.4     Snyder   column    -   Two   ball   micro   (Kontes
      K-569001-0219 or equivalent).
            4.2.3.5     Springs   -    1/2   inch   (Kontes  K-662750   or
      equivalent).
      4.2.4 Boiling chips  -  Solvent extracted with  methylene chloride,
approximately 10/40 mesh (silicon carbide or equivalent).
      4.2.5 pH meter - Capable of measuring to the nearest  0.01  units.
      4.2.6 Glass fiber filter paper - 1.2 ^m pore size  (Fisher Grade G4
or equivalent).
      4.2.7 Solid  sorbent  cartridges  -  Packed  with 2  g C18 (Baker or
equivalent).
      4.2.8 Vacuum manifold - Capable  of simultaneous extraction of up to
12 samples (Supelco or equivalent).
      4.2.9 Sample reservoirs - 60 ml  capacity (Supelco or  equivalent).
                             8315 - 5                         Revision 0
                                                          September 1994

-------
            4.2.10      Pipet  -   Capable  of  accurately  delivering  0.10  ml
      solution (Pipetman or equivalent).

            4.2.11      Water bath - Heated, with concentric ring cover, capable
      of temperature control (+ 2°C).   The bath  should  be  used  under  a  hood.

            4.2.12      Sample shaker  - Controlled temperature incubator (+ 2°C)
      with  orbital   shaking  (Lab-Line Orbit   Environ-Shaker  Model  3527  or
      equivalent).

            4.2.13      Syringes  -  5 mL,   500  juL,   100  /jL,  (Luer-Lok  or
      equivalent).

            4.2.14      Syringe  Filters  -   0.45 /urn  filtration disks  (Gelman
      Acrodisc 4438 or equivalent).

      4.3   Apparatus and Materials for Option 2

            4.3.1 Syringes  -  10 ml,  with  Luer-Lok  type   adapter,  used  to
      backflush the sample cartridges by gravity feed.

            4.3.2 Syringe Rack -  made  of an  aluminum plate with adjustable legs
      on all four corners.   Circular  holes  of  a diameter  slightly larger than
      the diameter of the 10 mL syringes are drilled through the plate to allow
      batch processing of cartridges for cleaning, coating, and sample elution.
      A plate (0.16 x 36 x 53 cm) with  45 holes  in a 5x9 matrix is recommended.
      See Figure 2 in Method 0100.

      4.4   Volumetric Flasks -  5 mL, 10 mL,  and 250 or 500 mL.

      4.5   Vials -  10  or 25 mL, glass with Teflon-lined screw  caps or crimp
tops.

      4.6   Balance - Analytical, capable of accurately weighing to 0.0001 g.

      4.7   Glass Funnels

      4.8   Polyethylene Gloves  - used to handle silica gel  cartridges.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used in all  tests.  Unless
otherwise  indicated,  it  is  intended  that   all  reagents  shall  conform  to the
specifications of the Committee on Analytical Reagents  of the American Chemical
Society, where  such  specifications  are available.  Other grades  may be used,
provided it is first,  ascertained  that the reagent is of  sufficiently high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water -  Water in which an interferant is not
observed at the method detection limit for  the compounds of interest.
                                   8315 - 6                         Revision 0
                                                                September 1994

-------
      5.3   Formalin - Solution of formaldehyde  (CH20) in organic-free  reagent
water, nominally 37.6 percent (w/w).  Exact concentration will  be determined  for
the stock solution in Sec. 5.7.1.1.

      5.4   Aldehydes and  ketones  -  analytical  grade,  used for  preparation  of
DNPH derivative standards of target analytes other than  formaldehyde.   Refer to
the target analyte list.

      5.5   Option 1 Reagents

            5.5.1   Methylene chloride,  CH2C12 -  HPLC grade or equivalent.

            5.5.2   Acetonitrile,  CH3CN   -  HPLC  grade or  equivalent.

            5.5.3   Sodium hydroxide  solutions,  NaOH,  1.0 N and 5 N.

            5.5.4   Sodium chloride,  NaCl,  saturated  solution -  Prepare   by
      dissolving an  excess of the  reagent grade solid in organic-free  reagent
      water.

            5.5.5   Sodium sulfite  solution, Na2S03, 0.1  M.

            5.5.6   Sodium sulfate,  Na2S04  - granular, anhydrous.

            5.5.7   Citric Acid,  C8H807,  1.0 M solution.

            5.5.8   Sodium Citrate, C6H5Na307.2H20, 1.0 M trisodium salt dihydrate
      solution.

            5.5.9   Acetic acid (glacial),  CH3C02H.

            5.5.10  Sodium acetate,  CH3C02Na.

            5.5.11  Hydrochloric  Acid, HC1, 0.1  N.

            5.5.12  Citrate buffer,  1 M,  pH 3  -  Prepare  by adding  80 ml of  1 M
      citric  acid  solution  to 20 ml  of  1  M  sodium  citrate  solution.    Mix
      thoroughly.  Adjust pH with NaOH or  HC1 as  needed.

            5.5.13  pH  5.0 Acetate  buffer (5M)   -  Formaldehyde analysis only.
      Prepared  by  adding  40  ml  5M  acetic acid  solution  to  60 ml  5M  sodium
      acetate solution.   Mix  thoroughly.   Adjust pH with  NaOH  or HC1 as  needed.

            5.5.14  2,4-Dinitrophenylhydrazine, 2,4-(02N)2C6H3]NHNH2, (DNPH),  70%
      in organic-free reagent water (w/w).

                    5.5.14.1   DNPH (3.00 mg/mL)  - Dissolve 428.7 mg of 70% (w/w)
            DNPH solution  in 100 ml acetonitrile.

            5.5.15  Extraction fluid for Option 1 -  Dilute 64.3 ml of 1.0 N NaOH
      and 5.7 ml glacial  acetic acid  to  900 ml with organic-free reagent water.
                                   8315 - 7                         Revision 0
                                                                September  1994

-------
Dilute to 1 liter with organic-free reagent  water.  The pH should be 4.93
± 0.02.

5.6   Option 2 Reagents

      5.6.1   Acetonitrile,  CH3CN  -  UV grade.

      5.6.2   2,4-Dinitrophenylhydrazine,  C6H6N404,  (DNPH) - recrystallize
at least twice with UV grade acetonitrile  using the procedure in Appendix
A.

5.7   Stock Standard Solutions for Option 1

      5.7.1   Stock formaldehyde  (approximately 1000  mg/L)  -  Prepare by
diluting  an   appropriate   amount  of   the   certified  or  standardized
formaldehyde  (approximately  265 jxL) to  100  ml with organic-free reagent
water.  If a certified formaldehyde solution is not available or there is
any question regarding the quality  of a certified solution, the solution
may be standardized using the procedure in Sec. 5.7.1.1.

              5.7.1.1    Standardization of formaldehyde stock solution -
      Transfer a 25 ml aliquot of a 0.1  M Na2S03 solution to a beaker and
      record  the  pH.   Add  a 25.0 ml aliquot of  the  formaldehyde stock
      solution (Sec. 5.18.1) and record the pH.  Titrate  this mixture back
      to the original  pH using 0.1 N HC1.  The formaldehyde concentration
      is calculated using the following equation:
              Concentration (mg/L)  =
      where:
             (30.03)(N HCl)(mL HC1)

                    25.0 ml
              N HC1
              mL HC1
              30.03
Normality of HC1  solution  used (in milli-
equivalents/mL) (1 mmole of HC1 = 1 mini-
equivalent of HC1)
mL of standardized HC1 solution used
Molecular  of weight  of formaldehyde  (in
mg/mmole)
      5.7.2   Stock  aldehyde(s)  and  ketone(s)  -  Prepare by  adding  an
appropriate  amount of  the  pure material  to  90 mL  of acetonitrile and
dilute to  100 mL,  to give a final concentration of 1000 mg/L.

5.8   Stock  Standard Solutions  for Option  2

      5.8.1   Preparation of the DNPH Derivatives for HPLC analysis

              5.8.1.1    To  a  portion  of  the recrystallized  DNPH, add
      sufficient  2N HC1 to obtain  an  approximately saturated  solution.
      Add  to this solution the target analyte  in molar excess of the DNPH.
      Filter the DNPH derivative precipitate,  wash it with 2N HC1,  wash  it
      again  with water,  and allow it to  dry in  air.
                             8315  - 8
                                Revision 0
                            September 1994

-------
                    5.8.1.2    Check the purity of the DNPH derivative by melting
            point determination or HPLC analysis.   If the impurity level is not
            acceptable, recrystallize  the derivative in  acetonitrile.   Repeat
            the purity check and recrystallization as necessary until 99% purity
            is achieved.

            5.8.2   Preparation  of DNPH  Derivative Standards and  Calibration
      Standards for HPLC analysis

                    5.8.2.1    Stock  Standard  Solutions  -  Prepare  individual
            stock  standard  solutions  for  each  of  the target  analyte  DNPH
            derivatives   by   dissolving   accurately   weighed   amounts   in
            acetonitrile.  Individual  stock  solutions of approximately 100 mg/L
            may be  prepared by  dissolving 0.010  g of the  solid  derivative in
            100 ml of acetonitrile.

                    5.8.2.2    Secondary  Dilution  Standard(s)   -   Using  the
            individual  stock  standard  solutions,  prepare  secondary  dilution
            standards in acetonitrile containing  the DNPH derivatives from the
            target  analytes mixed together.   Solutions of  100  ^g/i may  be
            prepared  by  placing 100  jiL of a  100  mg/L  solution  in a  100  mL
            volumetric flask and diluting to the  mark with  acetonitrile.

                    5.8.2.3    Calibration   Standards   -   Prepare   a   working
            calibration standard mix from the secondary dilution standard, using
            the mixture of  DNPH derivatives at concentrations of  0.5-2.0 /ig/L
            (which  spans  the  concentration  of interest for most  indoor  air
            work).  The concentration of the DNPH  derivative in the standard mix
            solutions may need to  be  adjusted to reflect relative concentration
            distribution in a real sample.

      5.9   Standard Storage  -  Store  all  standard  solutions  at  4°C  in  a glass
vial with a Teflon-lined  cap,  with minimum headspace,  and in the dark.   They
should be stable for about 6 weeks.  All  standards should be checked frequently
for signs  of  degradation or  evaporation,  especially just  prior  to preparing
calibration standards from them.

      5.10  Calibration Standards

            5.10.1  Prepare   calibration   solutions  at   a  minimum   of   5
      concentrations for each  analyte  of interest in organic-free reagent water
      (or acetonitrile  for  Option 2)  from  the  stock standard solution.   The
      lowest concentration of  each  analyte should be at, or just above, the MDLs
      listed in Tables  1  or 2.   The other concentrations of the  calibration
      curve should correspond  to the expected range of concentrations found in
      real samples.
                                   8315 - 9                         Revision 0
                                                                September 1994

-------
6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the introductory material to this Chapter, Organic Analytes, Sec.
4.1.

      6.2   Samples  must  be  refrigerated at  4°C.   Aqueous  samples must  be
derivatized and extracted within 3 days of sample collection.  The holding times
of  leachates  of solid samples should  be  kept  at a minimum.   All  derivatized
sample extracts should be analyzed within 3 days after preparation.

      6.3   Sample;; collected by Methods  0011  or 0100  must be  refrigerated at
4°C.  It  is recommended that samples be extracted and analyzed within 30 days of
collection.


7.0   PROCEDURE

      7.1   Extraction of Solid Samples (Option 1)

            7.1.1   All  solid  samples  should be made as homogeneous as possible
      by stirring and removal of sticks, rocks, and other extraneous material.
      When the sample is not dry,  determine the  dry weight  of the sample, using
      a  representative  aliquot.    If  particle  size  reduction  is  necessary,
      proceed as per Method 1311.

                    7.1.1.1    Determination of dry weight  -  In certain cases,
            sample results are desired based on a dry weight basis.  When such
            data are  desired  or  required, a portion of  sample for dry weight
            determination should be weighed out at the  same time as the portion
            used for analytical determination.

                    WARNING:   The drying oven should be contained in a hood or
                              vented.  Significant laboratory contamination may
                              result   from  drying  a   heavily  contaminated
                              hazardous waste sample.

                    7.1.1.2    Immediately  after  weighing  the   sample  for
            extraction,  weigh 5-10  g of  the  sample  into a  tared  crucible.
            Determine the % dry weight  of the sample  by  drying  overnight at
            105°C.   Allow to cool  in a desiccator before weighing:

                                    g  of dry sample
                    % dry weight  = 	  x 100
                                       g of sample

            7.1.2   Measure  25 g of  solid into a 500 mL  bottle with a Teflon
      lined screw  cap or crimp top, and  add 500 mL of extraction fluid (Sec.
      5.5.15).  Extract the  solid by rotating the bottle at approximately 30 rpm
      for  18  hours.   Filter the extract  through  glass  fiber  filter paper and
      store in  sealed bottles at 4°C.  Each mL  of  extract represents 0.050 g
      solid.     Smaller  quantities   of   solid  sample   may   be   used  with
                                   8315 - 10                        Revision 0
                                                                September  1994

-------
correspondingly reduced volumes of extraction fluid maintaining the 1:20
mass to volume ratio.

7.2   Cleanup and Separation (Option 1)

      7.2.1   Cleanup procedures may not  be necessary  for  a relatively
clean sample matrix.   The  cleanup  procedures  recommended in this method
have been used for the analysis  of various  sample types.  If particular
samples  demand the use of an alternative cleanup procedure, the analyst
must determine the elution profile and demonstrate  that the recovery of
formaldehyde from a spiked  sample is greater than 85%.   Recovery may be
lower for samples which form emulsions.

      7.2.2   If the sample is not clear, or the complexity is  unknown, the
entire sample should be centrifuged  at 2500 rpm  for 10 minutes.  Decant
the  supernatant  liquid from the  centrifuge bottle, and  filter through
glass fiber filter paper into a container which can be tightly sealed.

7.3   Derivatization (Option 1)

      7.3.1   For  aqueous samples,  measure an  aliquot of sample which is
appropriate  to  the  anticipated  analyte concentration  range (nominally
100 ml).   Quantitatively  transfer the sample  aliquot to  the  reaction
vessel (Sec. 4.2).

      7.3.2   For  solid samples, 1  to  10  ml of extract  (Sec.  7.1)  will
usually be  required.   The amount  used for  a particular  sample must be
determined through preliminary experiments.

      NOTE:   In cases  where the selected sample or extract volume is less
              than 100  ml, the total volume of the  aqueous layer  should be
              adjusted  to 100 ml with organic-free reagent water.  Record
              original  sample  volume prior to dilution.

      7.3.3   Derivatization  and  extraction  of the target analytes may be
accomplished using the liquid-solid  (Sec.  7.3.4)  or liquid-liquid (Sec.
7.3.5) procedures.

      7.3.4   Liquid-Solid  Derivatization  and  Extraction

              7.3.4.1    For analytes other than formaldehyde, add 4 ml of
      citrate buffer and adjust the pH to  3.0  +  0.1 with 6M HC1  or 6M
      NaOH.  Add  6 ml of DNPH  reagent,  seal  the container, and place in a
      heated (40°C),  orbital  shaker (Sec.  4.2.12) for 1 hour.  Adjust the
      agitation to produce a gentle swirling of the reaction solution.

              7.3.4.2   If formaldehyde is the only analyte of interest,
      add 4 mL acetate  buffer  and adjust pH to 5.0 +  0.1 with  6M  HC1 or 6M
      NaOH.  Add  6 ml of DNPH  reagent,  seal  the container, and place in a
      heated (40°C),  orbital  shaker (Sec.  4.2.12) for 1 hour.  Adjust the
      agitation to produce a gentle swirling of the reaction solution.
                             8315  -  11                         Revision 0
                                                          September 1994

-------
        7.3.4.3    Assemble the  vacuum manifold  and connect  to  a
water aspirator or vacuum  pump.   Attach a 2 g  sorbent cartridge  to
the vacuum  manifold.   Condition  each  cartridge by passing  10  ml
dilute citrate buffer (10  ml  of  1  M citrate buffer dissolved in 250
ml of organic-free reagent water) through each sorbent cartridge.

        7.3.4.4    Remove  the  reaction  vessel  from  the  shaker
immediately at the end of  the one hour reaction period and  add 10 ml
saturated NaCl solution to the vessel.

        7.3.4.5    Quantitatively transfer the reaction solution to
the sorbent cartridge  and apply  a vacuum so  that  the  solution is
drawn through the cartridge  at  a  rate  of 3 to 5 mL/min.   Continue
applying the vacuum for about 1 minute after the liquid sample has
passed through the cartridge.

        7.3.4.6    While maintaining the vacuum conditions described
in Sec.  7.3.4.4,  elute  each cartridge  train with approximately 9 ml
of acetonitrile  directly  into a 10 ml volumetric flask.  Dilute the
solution to volume with acetonitrile, mix thoroughly, and place in
a tightly sealed vial until  analyzed.
        NOTE:
                  Because this method uses an excess  of DNPH,  the
                  cartridges  will   remain  a  yellow  color  after
                  completion of Sec.  7.3.4.5.  The presence of this
                  color  is  not  indicative  of   the   loss  of  the
                  analyte derivatives.

7.3.5   Liquid-Liquid Derivatization  and  Extraction

        7.3.5.1    For analytes other than formaldehyde,  add 4 mL of
citrate buffer  and  adjust  the pH to 3.0 + 0.1  with  6M HC1  or 6M
NaOH.  Add 6 mL  of DNPH  reagent,  seal the container, and place in a
heated (40°C),  orbital  shaker for 1  hour.  Adjust the agitation to
produce a gentle swirling of the reaction solution.

        7.3.5.2    If formaldehyde is the only analyte of interest,
add 4 mL acetate buffer  and  adjust pH to 5.0 ± 0.1 with 6M HC1 or 6M
NaOH.  Add 6 mL  of DNPH  reagent,  seal the container, and place in a
heated (40°C),  orbital  shaker for 1  hour.  Adjust the agitation to
produce a gentle swirling of the reaction solution.

        7.3.5.3    Serially  extract  the  solution with three  20 mL
portions of methylene chloride using a 250 mL  separatory  funnel.  If
an emulsion  forms upon  extraction,  remove  the entire emulsion and
centrifuge  at  2000  rpm for  10 minutes.   Separate the  layers  and
proceed with the  next extraction.   Combine the  methylene chloride
layers  in  a 125  mL  Erlenmeyer flask  containing  5.0 grams  of
anhydrous sodium  sulfate.   Swirl  contents  to complete the extract
drying process.
                       8315 - 12                        Revision 0
                                                    September 1994

-------
        7.3.5.4    Assemble a Kuderna-Danish  (K-D)  concentrator by
attaching a 10 ml concentrator tube  to  a  500 ml evaporator flask.
Pour the extract  into the evaporator flask  being  careful to minimize
transfer of sodium sulfate granules.  Wash  the Erlenmeyer flask with
30 ml of methylene  chloride  and add wash to the evaporator flask to
complete quantitative transfer.

        7.3.5.5    Add  one  to  two   clean  boiling  chips  to  the
evaporative flask and attach a three ball Snyder  column.  Prewet the
Snyder column by adding about 1 ml methylene chloride to the top.
Place the K-D apparatus on  a hot  water  bath  (80-90°C) so that the
concentrator tube is partially  immersed in the hot water and the
entire lower rounded surface of the flask  is  bathed with hot vapor.
Adjust  the  vertical  position  of  the  apparatus  and  the  water
temperature, as  required,  to  complete the concentration  in 10-15
min.  At  the  proper rate  of distillation the balls of  the column
will  actively chatter,  but  the  chambers  will  not flood  with
condensed solvent.   When the apparent volume of liquid  reaches 5 ml,
remove the K-D apparatus and allow it to  drain and cool for at least
10 min.

        7.3.5.6    Prior to  liquid  chromatographic  analysis,  the
extract solvent must be exchanged to acetonitrile.  The analyst must
ensure  quantitative transfer  of the  extract  concentrate.    The
exchange is performed as follows:

            7.3.5.6.1   Remove  the  three-ball  Snyder column  and
        evaporator  flask.   Add  5 ml of acetonitrile  , a  new glass
        bead or boiling chip, and attach the micro-Snyder column to
        the  concentrator tube.   Concentrate the  extract using 1 ml
        of acetonitrile  to prewet the Snyder  column.   Place the K-D
        apparatus on the water bath so that  the concentrator tube is
        partially  immersed  in  the hot water.  Adjust  the vertical
        position  of the  apparatus and  the water temperature,  as
        required, to complete concentration.  At the proper rate of
        distillation the balls of the column will actively chatter,
        but  the chambers will not  flood.   When the  apparent volume
        of liquid reaches  less than  5 ml,  remove the  K-D apparatus
        and  allow it to drain and  cool for  at least 10 minutes.

            7.3.5.6.2   Remove the Snyder column and rinse the flask
        and  its lower joint with 1-2 ml of acetonitrile  and add to
        concentrator tube.  Quantitatively transfer the sample to a
        10 mL volumetric flask using a 5 ml  syringe with an attached
        Acrodisc 0.45 /im filter cassette.   Adjust the extract volume
        to 10  ml.   Stopper the  flask and store refrigerated at 4°C
        if further processing will  not be performed immediately.  If
        the  extract  will  be  stored longer than two   (2)  days,  it
        should be  transferred  to  a vial  with  a  Teflon lined screw
        cap  or  crimp  top.    Proceed  with  HPLC  chromatographic
        analysis  if  further  cleanup is not  required.
                       8315  -  13                         Revision 0
                                                    September 1994

-------
7.4   Extraction of Samples from Methods 0011 and 0100 (Options 1 and 2)

      7.4.1   Stack  gas  samples  collected by  Method 0011  (Option  1)

              7.4.1.1    Measure the volume of the aqueous phase  of the
      sample prior to extraction (for moisture determination in case the
      volume was  not  measured  in  the  field).   Pour  the sample  into  a
      separatory funnel  and drain the methylene chloride into a volumetric
      flask.

              7.4.1.2    Extract the aqueous  solution  with two  or three
      aliquots of methylene chloride.  Add the methylene chloride extracts
      to the volumetric flask.

              7.4.1.3    Fill   the  volumetric  flask   to  the  line  with
      methylene chloride.  Mix well and remove an aliquot.

              7.4.1.4    If  high   concentrations  of   formaldehyde  are
      present, the extract can be diluted with mobile phase, otherwise the
      extract solvent must  be  exchanged as described in Sec. 7.3.5.5.  If
      low concentrations of formaldehyde are present, the  sample should be
      concentrated during the solvent exchange procedure.

              7.4.1.5    Store the  sample at 4°C.   If the  extract will be
      stored longer  than two days,  it should be transferred to a vial with
      a Teflon-lined screw cap,  or  a crimp top with a Teflon-lined septum.
      Proceed with HPLC  chromatographic analysis if further cleanup  is not
      required.

      7.4.2   Ambient  air samples  collected by Method  0100  (Option  2)

              7.4.2.1    The samples will  be received by the laboratory in
      a friction-top  can containing 2  to 5 cm of granular charcoal, and
      should  be  stored  in  this can, in  a  refrigerator,  until  analysis.
      Alternatively,  the samples  may  also   be  stored  alone   in  their
      individual glass containers.  The time between  sampling and analysis
      should not exceed 30 days.

              7.4.2.2    Remove  the sample  cartridge  from the  labeled
      culture  tube.   Connect the  sample  cartridge  (outlet or  long end
      during sampling) to a clean  syringe.

              NOTE:      The liquid flow during desorption  should  be in
                        the opposite direction from the  air flow during
                        sample collection (i.e, backflush the cartridge).

              7.4.2.3    Place the  cartridge/syringe in the syringe rack.

              7.4.2.4    Backflush the cartridge (gravity  feed) by passing
      6 ml  of acetonitrile  from the syringe  through  the cartridge to a
      graduated test  tube,  or to a 5 ml volumetric flask.
                             8315 - 14                        Revision 0
                                                          September 1994

-------
              NOTE:      A dry cartridge has  an acetonitrile holdup volume
                        slightly greater than 1 ml.  The eluate flow may
                        stop before  the acetonitrile in the  syringe is
                        completely drained into the cartridge because of
                        air trapped between the cartridge filter and the
                        syringe Luer-Lok tip.  If this happens, displace
                        the  trapped  air  with  the  acetonitrile  in  the
                        syringe  using  a  long-tip  disposable  Pasteur
                        pipet.

              7.4.2.5    Dilute to the 5 ml mark with acetonitrile.  Label
      the  flask  with  sample  identification.   Pipet two  aliquots  into
      sample vials having Teflon-lined septa.

              7.4.2.6    Store  the sample  at  4°C.    Proceed  with  HPLC
      chromatographic analysis of the first  aliquot if further cleanup is
      not required.  Store  the  second  aliquot  in the refrigerator until
      the results of the  analysis  of the  first  aliquot  are complete and
      validated.     The second  aliquot  can be   used  for  confirmatory
      analysis, if necessary.

7.5   Chromatographic Conditions (Recommended):

      7.5.1   Option 1  -  For aqueous  samples, soil or waste samples, and
stack gas samples collected by Method 0011.

      Column:                 CIS, 4.6 mm x  250 mm ID, 5 /zm particle size
      Mobile Phase Gradient:  70%/30% acetonitrile/water (v/v), hold for
                              20 min.
                              70%/30%    acetonitrile/water   to    100%
                              acetonitrile in 15 min.
                              100% acetonitrile for 15 min.
      Flow Rate:               1.2 mL/min
      Detector:               Ultraviolet, operated at 360 nm
      Injection Volume:       20 fj,l

      7.5.2   Option 2  - For ambient air samples collected by Method 0100.

      Column:                 Two  HPLC columns,  4.6 mm  x  250 mm  ID,
                              (Zorbax ODS, or equivalent) in series
      Mobile Phase Gradient:  60%/40% CH3CN/H20,  hold  for 0  min.
                              60%/40% to 75%/25%  CH3CN/H20,  linearly in 30
                              min.
                              75%/25% to 100%/0%  CH3CN/H20,  linearly in 20
                              min.
                              100% CH3CN for 5  minutes.
                              100%/0% to 60%/40% CH3CN/H20, linearly in 1
                              min.
                              60%/40% CH3CN/H20 for 15 minutes.
      Detector:               Ultraviolet, operated at 360 nm
      Flow Rate:               1.0 mL/min
      Sample Injection volume:25 ^l  (suggested)


                             8315  - 15                         Revision 0
                                                          September 1994

-------
      NOTE:   For  Options  1 and 2, analysts are  advised  to adjust their
              HPLC  systems  to  optimize  chromatographic  conditions  for
              their  particular  analytical  needs.    The   separation  of
              acrolein,  acetone,  and  propionaldehyde should be a minimum
              criterion  of the  optimization  in Option 2.

      7.5.3   Filter and degas the mobile phase to remove dissolved gasses,
using the following procedure:

              7.5.3.1    Filter  each   solvent  (water  and  acetonitrile)
      through a 0.22  pm polyester membrane filter, in an  all  glass and
      Teflon suction filtration apparatus.

              7.5.3.2    Degas  each  filtered  solution  by  purging  with
      helium for 10-15 minutes  (100 mL/min) or by heating  to 60°C for 5-10
      minutes in  an  Erlenmeyer  flask covered  with a watch  glass.   A
      constant back pressure restrictor (350 kPa) or 15-30 cm of 0.25 mm
      ID Teflon tubing  should  be  placed  after the  detector to eliminate
      further mobile phase outgassing.

              7.5.3.3    Place  the  mobile  phase  components  in  their
      respective HPLC solvent reservoirs,  and program the gradient system
      according to the conditions listed  in  Sec.  7.5.2.  Allow the system
      to pump for  20-30 minutes at a  flow  rate of 1.0  mL/min  with the
      initial solvent  mixture  ratio  (60%/40% CH3CN/H20).   Display the
      detector output on a strip chart recorder or similar output device
      to establish a stable baseline.

7.6   Calibration

      7.6.1   Establish  liquid  chromatographic  operating  conditions  to
produce  a  retention  time similar to  that indicated in Table  1  for the
liquid-solid derivatization and extraction or in  Table 2 for liquid-liquid
derivatization  and extraction.    For determination  of  retention  time
windows,  see  Sec.  7.5  of Method   8000.    Suggested  chromatographic
conditions are provided in Sec. 7.5.

      7.6.2   Process    each   calibration   standard  solution   through
derivatization  and extraction,  using the  same procedure  employed for
sample processing (Sees. 7.3.4 or 7.3.5).

      7.6.3   Analyze  a  solvent blank to  ensure  that the  system is clean
and interference free.

      NOTE:   The  samples  and standards must be allowed to come to ambient
              temperature  before  analysis.

      7.6.4   Analyze  each  processed calibration  standard  using  the
chromatographic conditions listed in  Sec.  7.5, and tabulate  peak area
against  calibration solution concentration  in /jg/L.
                             8315 - 16                        Revision 0
                                                          September 1994

-------
      7.6.5   Tabulate  the peak  area  along with  standard  concentration
injected to determine  the response factor (RF) for the  analyte at each
concentration  (see  Sec.  7.8.1  for  equations).    The percent  relative
standard deviation  (%RSD) of the  mean  RF of  the  calibration standards
should be no greater than ± 20 percent or a system check will have to be
performed.   If a calibration  check after  the  system  check  does not meet
the  criteria, a  recalibration  will   have to be  performed.    If  the
recalibration does  not meet  the  established   criteria,  new  calibration
standards must be made.

      7.6.6   The  working calibration curve must  be  verified  each day,
before  and after  analyses  are  performed,  by  analyzing  one  or  more
calibration standards.   The  response obtained should fall  within  + 15
percent of the initially established response or a system check will have
to be performed.   If a calibration check after the system check does not
meet the criteria, the system must be recalibrated.

      7.6.7   After  10  sample  runs,  or  less, one  of  the  calibration
standards must be reanalyzed to ensure that the DNPH derivative response
factors remain within +15% of the original calibration response factors.

7.7   Sample Analysis

      7.7.1   Analyze samples by HPLC, using conditions established  in Sec.
7.5.  For  analytes  to  be analyzed by Option 1, Tables 1 and 2 list the
retention times and MDLs that were obtained under these conditions.  For
Option 2 analytes, refer to Figure 3 for the sample chromatogram.

      7.7.2   If the peak  area exceeds  the  linear range of the calibration
curve, a smaller sample injection volume should be used.   Alternatively,
the final solution may be diluted with acetonitrile and reanalyzed.

      7.7.3   After   elution   of  the   target   analytes,   calculate  the
concentration of analytes found in the samples using the equations found
in Sec.  7.8 or the specific sampling method used.

      7.7.4   If the peak  area measurement  is prevented by the presence of
observed interferences, further cleanup is required.

7.8   Calculations

      7.8.1   Calculate each  response factor,   mean response  factor,  and
percent relative standard deviation as follows:

               Concentration  of standard  injected,
                        Area of signal
                            8315  -  17                         Revision 0
                                                          September 1994

-------
                _         IRF(
      Mean RF = RF      = 	
                !  I (RF, - RF)2 /N-l
      %RSD    =  	—	  x 100%
                        RF

where:

      RF      =    Mean  response  factor or mean of  the response factors
                  using the 5 calibration concentrations.
      RFj      =    Response factor for calibration standard i  (i =  1-5).
      %RSD    =    Percent  relative  standard  deviation of  the response
                  factors.
      N       =    Number of calibration  standards.

      7.8.2   Calculate the analyte concentrations  in liquid samples  as
follows:

      Concentration of  aldehydes in fj.g/1 -  (RF)(Area  of signal)(100/V8)

where:

      RF      =    Mean  response  factor for  a  particular analyte.
      Vs      =    Number  of ml of sample (unitless).

      7.8.3   Calculate  the  analyte concentration  in solid  samples  as
follows:

      Concentration of  aldehydes  in ng/g = (RF)(Area  of signal)(20/ Vex)

where:

      RF      =    Mean  response  factor for  a  particular analyte.
      Vex     =    Number of ml extraction  fluid aliquot  (unitless).

      7.8.4   Calculate  the  concentration of  formaldehyde  in  stack  gas
samples  (Method 0011)  as follows:  (Option  1)
                             8315  -  18                         Revision 0
                                                           September 1994

-------
                    7.8.4.1   Calculation of Total  Formaldehyde:   To determine
            the total formaldehyde  in mg, use the  following  equation:

                                         [g/mole  formaldehyde]
Total mg formaldehyde = Cd x V x DF x 	 x  10~3  mg/Mg
                                        [g/mole DNPH derivative]

            where:

                    Cd   =     measured   concentration   of  DNPH-formaldehyde
                              derivative, mg/L
                    V   =     organic extract volume,  ml
                    DF   =     dilution  factor

                    7.8.4.2   Formaldehyde concentration in stack gas: Determine
            the formaldehyde concentration in the stack gas using the following
            equation:

                    Cf = K  [total formaldehyde, mg]  / Vm(8td)

            where:

                    K         =     35.31  ft3/m3,   if   Vm(8td)  is  expressed  in
                                    English units
                                    1.00 m3/m3,  if Vm(std) is expressed in metric
                                    units
                    Vm(std)      =     volume of gas sample as measured by dry gas
                                    meter,  corrected  to  standard  conditions,
                                    dscm (dscf)

            7.8.5   Calculation of the Concentration of Formaldehyde and Other
      Carbonyls from Indoor Air Sampling by Method  0100.  (Option  2)

                    7.8.5.1   The concentration of  target analyte  "a" in air at
            standard conditions  (25°C  and  101.3  kPa),   Concastd  in  ng/L,  may be
            calculated using the following equation:

                           (AreaJ (RF) (VolJ (MWJ (1000 ng/Mg)
                    Conca = 	  x DF
                               (MWd)(VTotStd)(1000 ml/I)
            where:
                    Areaa      =     Area of  the  sample  peak for  analyte "a"
                    RF        =     Mean response  factor for analyte  "a"  from
                                    the calibration  in  M9/L- (See Sec, 7.8.1)
                    Vola       =     Total volume of the sample cartridge eluate
                                    (ml)
                    MWa        =     Molecular weight  of analyte  "a"  in g/mole
                    MWd        =     Molecular weight  of the DNPH derivative  of
                                    analyte  "a"  in g/mole
                                   8315 - 19                         Revision 0
                                                                 September 1994

-------
                "TolStd
                    DF
Total volume  of air sampled  converted to
standard  conditions  in  liters  (L).  (To
calculate  the  concentration  at  sampling
conditions  use  Vtot.)(See  Sec.  9.1.3  of
Method 0100)
Dilution  Factor  for the sample  cartridge
eluate,  if any.   If there  is  no dilution,
DF = 1
                    7.8.5.2    The target analyte "a" concentration at standard
            conditions may be converted  to parts per billion by volume,  Cone, in
            ppbv, using the following equation:
                    Conca in ppbv
            where:
                    Conca
                    22.4

                    MW,
  (Cone.) (22.4)

      (MWJ
Concentration of analyte "a" in ng/L
Ideal gas  law  volume (22.4 nl_ of  gas  = 1
nmole at standard conditions)
Molecular weight of  analyte "a"  in g/mole
(or ng/nmole)
8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and  Method  8000 for specific quality control
procedures.   Refer  to Table  4  for  QC  acceptance limits  derived  from  the
interlaboratory method validation study on Method 8315.


9.0   METHOD PERFORMANCE

      9.1   The MDLs for Option  1  listed in Table 1 were obtained using organic-
free reagent water and  liquid-solid extraction.  The MDLs for Option  1  listed in
Table 2 were obtained  using  organic-free  reagent water and methylene chloride
extraction.  Results reported in  Tables  1  and 2 were achieved using fortified
reagent water volumes of 100 mL.  Lower detection limits may be obtained using
larger sample volumes.

            9.1.1   Option  1 of this  method  has  been  tested  for  linearity of
      recovery from spiked organic-free reagent water and has been demonstrated
      to be applicable over the range 50-1000 jug/L  .

            9.1.2   To  generate the MDL and precision and accuracy data reported
      in this section,  analytes were segregated  into two spiking groups, A and
      B.   Representative  chromatograms  using  liquid-solid  and  liquid-liquid
      extraction  are  presented   in  Figures   1  (a  and b)  and 2  (a  and  b),
      respectively.
                                   8315 - 20
                                Revision 0
                            September  1994

-------
      9.2   The Sensitivity of Option 2 sampling (Method 0100) and analysis is
listed in Table 3.

      9.3   Method 8315, Option 1, was tested by 12 laboratories using reagent
water and ground waters spiked at six concentration levels over the range 30-2200
jug/L.  Method  accuracy  and  precision  were  found to be directly related to the
concentration of the analyte and independent of the sample matrix.  Mean recovery
weighted  linear   regression  equations,  calculated  as  a  function  of  spike
concentration, as well  as  overall  and  single-analyst  precision  regression
equations, calculated as functions of mean recovery, are presented in Table 5.
These equations  can  be  used  to estimate  mean  recovery and  precision  at  any
concentration value within the range tested.


10.0  REFERENCES

1.    "OSHA Safety and  Health Standards,  General Industry",  (29CRF1910).
      Occupational  Safety  and  Health  Administration,  OSHA  2206,  (Revised,
      January 1976).


11.0  SAFETY

      11.1  The toxicity or carcinogenicity of each  reagent  used in this method
has not been precisely defined; however,  each chemical compound should be treated
as a potential health hazard.   From this viewpoint,  exposure to these chemicals
must be reduced to the lowest possible level by whatever means available.  The
laboratory  is  responsible  for  maintaining a current  awareness file  of OSHA
regulations  regarding  the  safe  handling  of the chemicals specified  in this
method.    A  reference  file of material  safety data  sheets  should  also  be made
available to  all   personnel   involved  in  the chemical  analysis.    Additional
references to laboratory safety are available.

      11.2  Formaldehyde has been tentatively classified as a known or suspected,
human or mammalian carcinogen.
                                   8315  -  21                         Revision 0
                                                                September 1994

-------
                        TABLE 1.

        OPTION 1 - METHOD DETECTION LIMITS8 USING
                 LIQUID-SOLID EXTRACTION
Analyte             Retention Time                 MDL
                      (minutes)
Formaldehyde
Acetaldehyde
Propanal
Crotonaldehyde
Butanal
Cyclohexanone
Pentanal
Hexanal
Heptanal
Octanal
Nonanal
Decanal
5.3
7.4
11.7
16.1
18.1
27.6
28.4
34.1
35.0
40.1
40.4
44.1
6.2
43. 7b
11.0
5.9
6.3
5.8
15.3
10.7
10.0
6.9
13.6
4.4
 The  method  detection  limit  (MDL)  is  defined  as  the  minimum
 concentration  that  can be  measured with  99%  confidence that  the
 value   is   above  background  level.     With   the  exception   of
 acetaldehyde,  all reported  MDLs  are based upon analyses  of  6 to 8
 replicate  blanks spiked  at  25  ng/i.    The  MDL  was  computed  as
 fol1ows:

        MDL = t(N.1i0.01)(Std.  Dev.)

 where:

        t(N-i,o.oi>   =     Tne  upper  first  percentile  point   of  the
                         t-distribution with n-1 degrees of freedom.
         Std. Dev. =     Standard  deviation, calculated   using  n-1
                         degrees of  freedom.

 The reported MDL is  based upon analyses  of 3  replicate,  fortified
 blanks at 250
                        8315 - 22                         Revision 0
                                                      September 1994

-------
                             TABLE  2.

             OPTION 1 - METHOD DETECTION LIMITS8 USING
                     LIQUID-LIQUID  EXTRACTION
Analyte Retention Time
(minutes)
Formaldehyde
Acetaldehyde
Propanal
Crotonaldehyde
Butanal
Cyclohexanone
Pentanal
Hexanal
Heptanal
Octanal
Nonanal
Decanal
5.3
7.4
11.7
16.1
18.1
27.6
28.4
34.1
35.0
40.1
40.4
44.1
MDL
(M9/L)a
23.2
110. 2b
8.4
5.9
7.8
6.9
13.4
12.4
6.6
9.9
7.4
13.1
     The  method  detection  limit   (MDL)   is   defined   as  the  minimum
     concentration that can be  measured with 99% confidence that the value
     is above background level.  With  the  exception  of  acetaldehyde,  all
     reported MDLs  are based  upon  analyses of 6  to 8  replicate blanks
     spiked at 25 /ag/L.  The MDL was computed  as  follows:
            =     t(N.1i0.ol)(Std.  Dev.)

where:

     t
-------
                                   TABLE 3.

        OPTION 2 - SENSITIVITY (ppb, v/v) OF SAMPLING AND ANALYSIS FOR
        CARBONYL COMPOUNDS IN AMBIENT AIR USING AN ADSORBENT CARTRIDGE
                          FOLLOWED BY GRADIENT HPLC"
Compound
10
     Sample Volume (L)b

20    30    40    50   100   200   300   400   500
Acetaldehyde
Acetone
Acrolein
Benzaldehyde
Butyraldehyde
Crotonaldehyde
2,5-Dimethyl-
benzaldehyde
Formaldehyde
Hexanal
Isovaleraldehyde
Propionaldehyde
m-Tolualdehyde
o-Tolualdehyde
p-Tolualdehyde
Valeraldehyde
1
1
1
1
1
1

0
1
1
1
1
1
1
1
1
.36
.28
.29
.07
.21
.22

.97
.45
.09
.15
.28
.02
.02
.02
.15
0.68
0.64
0.65
0.53
0.61
0.61

0.49
0.73
0.55
0.57
0.64
0.51
0.51
0.51
0.57
0.45
0.43
0.43
0.36
0.40
0.41

0.32
0.48
0.36
0.38
0.43
0.34
0.34
0.34
0.38
0.34
0.32
0.32
0.27
0.30
0.31

0.24
0.36
0.27
0.29
0.32
0.25
0.25
0.25
0.29
0.27
0.26
0.26
0.21
0.24
0.24

0.19
0.29
0.22
0.23
0.26
0.20
0.20
0.20
0.23
0.14
0.13
0.13
0.11
0.12
0.12

0.10
0.15
0.11
0.11
0.13
0.10
0.10
0.10
0.11
0.07
0.06
0.06
0.05
0.06
0.06

0.05
0.07
0.05
0.06
0.06
0.05
0.05
0.05
0.06
0.05
0.04
0.04
0.04
0.04
0.04

0.03
0.05
0.04
0.04
0.04
0.03
0.03
0.03
0.04
0.03
0.03
0.03
0.03
0.03
0.03

0.02
0.04
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.03
0.02
0.02
0.02

0.02
0.03
0.02
0.02
0.03
0.02
0.02
0.02
0.02
   a  The ppb  values  are measured at 1 atm  and  25°C.  The  sample cartridge is
      eluted with 5 mL acetonitrile  and 25 /iL is injected  into the HPLC.  The
      maximum sampling flow through a DNPH-coated Sep-Pak is about 1.5 L/minute.

   b  A  sample volume  of  1000  L  was  also  analyzed.    The  results  show  a
      sensitivity of 0.01 ppb for all the target analytes.
                                   8315 - 24
                                                 Revision 0
                                             September 1994

-------
                             TABLE 4.

         PERFORMANCE-BASED QC ACCEPTANCE  LIMITS  CALCULATED
                USING THE COLLABORATIVE STUDY DATA
Spike
Analyte Concentration"
Formaldehyde
Propanal
Crotonaldehyde
Butanal
Cyclohexanone
Hexanal
Octanal
Decanal
160
160
160
160
160
160
160
160
X"
154
148
160
151
169
151
145
153
s»°
30.5
22.4
34.8
22.7
39.2
34.6
40.1
40.0
Acceptance
Limits, %d
39-153
50-134
35-165
52-137
32-179
30-159
15-166
21-171
Spike concentration, M9/L-
Mean recovery calculated  using the  reagent  water,  mean recovery,  linear
regression equation, ng/l.
Overall standard  deviation  calculated using the reagent  water,  overall
standard deviation linear regression equation,  M9/L.
Acceptance limits calculated as (X + 3sR)100/spike  concentration.
                            8315  - 25                         Revision 0
                                                          September 1994

-------
                                  TABLE  5.

WEIGHTED LINEAR REGRESSION EQUATIONS FOR MEAN RECOVERY AND PRECISION
Analyte
Formaldehyde
Propanal
Crotonaldehyde
Butanal
Cyclohexanone
Hexanal
Octanal
Decanal
Appl icable
Cone. Range
39.2-2450
31.9-2000
32.4-2030
35.4-2220
31.6-1970
34.1-2130
32.9-2050
33.2-2080
Reagent Water
X 0.909C + 8.79
SR 0.185X + 1.98a
sr 0.093X + 5.79
X 0.858C + 10.49
SR 0.140X + 1.63
sr 0.056X + 2.76
X 0.975C + 4.36
SR 0.185X + 5.15
sr 0.096X + 1.85
X 0.902C + 6.65
SR 0.149X + 0.21
s, 0.086X - 0.71
X 0.962C + 14.97
SR 0.204X + 4.73a
sr 0.187X + 3.46
X 0.844C + 15.81
SR 0.169X + 9.07
sr 0.098X + 0.37a
X 0.856C + 7.88
SR 0.200X + 11.17
sr 0.092X + 1.71a
X 0.883C + 12.00
SR 0.225X + 5.52
sr 0.088X + 2.28a
a Variance is not constant over concentration range.
X Mean recovery, M9/U exclusive of outliers.
SR Overall standard deviation, M9/L> exclusive of outl
sr Single-analyst standard deviation, M9/U exclusive
Ground Water
0.870C +14.84
0.177X + 13.85
0.108X + 6.24
0.892C + 22.22
0.180X + 12.37
0.146X + 2.08a
0.971C + 2.94
0.157X + 6.09
0.119X - 2.27
0.925C + 12.71
0.140X + 6.89
0.108X - 1.63a
0.946C + 28.95
0.345X + 5.02
0.123X + 7.64
0.926C + 9.16
0.132X + 8.31
0.074X - 0.40a
0.914C + 13.09
0.097X + 12.41
0.039X + 1.14
0.908C + 6.46
0.153X + 2.23
0.052X + 0.37
iers.
of outliers.
                                  8315  -  26
    Revision 0
September 1994

-------
                                    FIGURE la.

    OPTION 2 - LIQUID-SOLID PROCEDURAL STANDARD  OF  GROUP A ANALYTES AT 625 M9/L
  -0.80-
  -1.00-
 •-1.20-
V*

X
 -1.40-
 -1.80-
 -1.80-'
 -2.M
8

                1.00
  a.oo            3.00
         x 10* Minute*
4.00
                           Retention  Time
                             (minutes)
                                5.33
                               11.68
                               18.13
                               27.93
                               36.60
                               42.99
                     Analyte
                    Derivative
                   Formaldehyde
                   Propanal
                   Butanal
                   Cyclohexanone
                   Heptanal
                   Nonanal
                                     8315 - 27
                                       Revision 0
                                   September 1994

-------
                             FIGURE Ib.

  OPTION 1  - LIQUID-SOLID PROCEDURAL STANDARD OF GROUP B ANALYTES AT 625 M9/L
-0.00-
-0.80-
                                                       ^
                                                       Li
         1.00
2.00
3.00
4.00
                                  10*
                     Retention Time
                       (minutes)
                         7.50
                         16.68
                         26.88
                         32.53
                         40.36
                         45.49
                  Analyte
                  Derivative
                 Acetaldehyde
                 Crotonaldehyde
                 Pentanal
                 Hexanal
                 Octanal
                 Decanal
                              8315 - 28
                                 Revision 0
                              September 1994

-------
                                   FIGURE 2a.

  OPTION 1 - LIQUID-LIQUID PROCEDURAL STANDARD OF GROUP A ANALYTES AT 625 M9/L
-1.40-
             1.00
a.oo
          3.00
> 10* BinutM
4.00
                         Retention Time
                            (minutes)
                              5.82
                              13.23
                              20.83
                              29.95
                              37.77
                              43.80
                     Analyte
                    Derivative
                   Formaldehyde
                   Propanal
                   Butanal
                   Cyclohexanone
                   Heptanal
                   Nonanal
                                    8315 - 29
                                       Revision  0
                                   September 1994

-------
                                   FIGURE 2b.

  OPTION 1 - LIQUID-LIQUID PROCEDURAL STANDARD OF GROUP B ANALYTES AT 625
-8.00-
           1.00
f.OO
                                             3.00
                                       a 10*  •inutM
                                  4.00
                         Retention Time
                            (minutes)
                              7.79
                              17.38
                              27.22
                              32.76
                              40.51
                              45.62
                       Analyte
                      Derivative
                     Acetaldehyde
                     Crotonaldehyde
                     Pentanal
                     Hexanal
                     Octanal
                     Decanal
                                    8315 - 30
                                         Revision 0
                                     September 1994

-------
                         FIGURE 3.
OPTION 2  - CHROMATOGRAPHIC  SEPARATION OF THE  DNPH DERIVATIVES
                  OF 15 CARBONYL COMPOUNDS
          ONPH
                      JuJ
                                                 14
                                                  15
             10
                            20

                                  TIME, mm

                     Peak Identification
            30
                           40
   Number   Compound
Concentration^/ L)
     1      Formaldehyde              1.140
     2      Acetaldehyde              1.000
     3      Acrolein                 1.000
     4      Acetone                  1.000
     5      Propanal                 1.000
     6      Crotonaldehyde            1.000
     7      Butanal                  0.905
     8      Benzaldehyde              1.000
     9      Isovaleraldehyde          0.450
    10      Pentanal                 0.485
    11      o-Tolualdehyde            0.515
    12      m-Tolualdehyde            0.505
    13      p-Tolualdehyde            0.510
    14      Hexanal                  1.000
    15      2,4-Dimethylbenzaldehyde  0.510
                        8315 - 31
                         Revision 0
                     September 1994

-------
                                       METHOD 8315

                       DETERMINATION OF CARBONYL COMPOUNDS
              BY HIGH  PERFORMANCE  LIQUID  CHROMATOGRAPHY  (HPLC)
                                          Start
          0
Ambient Air (Option 2)
7.0 What is
the sample
  marix?
stack Gas (Option i
                                            (Media (Option 1)
  7.1.1-7.1.1.1
Homogenize sample
 and determine dry
     weight
      Solid
   7.1.2 Extract
   sample tor 18
  hours; filter and
   store extract
7.0 Is media
  solid or
 aqueous?
                                            I Aqueous
                       Is sample
                     dear or sample
                       complexity
                        known?
                                            Yes
                               7.2.2 Centrifuge sample
                                 at 2500 rpm for 10
                                  minutes; decant
                                     and filter
 7.3.2 Measure 1-10
  ml extract; adjust
  volume to 100 mL
     with water
        Solid
  7.3.1 Is
media solid
                                         Aqueous
                    7.3.1 Measure
                   aliquot of sample;
                   adjust volume to
                   100 ml with water
                                      7.3.S.S Exchange
                                     solvent to methanol
                                         O
                                       8315  -  32
                                                                  Revision  0
                                                            September  1994

-------
                                       METHOD  8315
                                         continued
                                                  7.4.1.1 Measure volume
                                                   of aqueous phase of
                                                 sample: pour sample Into
                                                   separatory funnel and
                                                 drain methytene chloride
                                                  (from Method 0011) Into
                                                     volumetric flask
                                                          I
                                                  7.4.1.2 Extract aqueous
                                                  solution with methvtene
                                                  chloride; add methytene
                                                    chloride extracts to
                                                      volumetric task
                                                          I
                                                  7.4.1.3 Mute to volume
                                                  w«h metiytene chloride;
                                                  mix wed; remove aliquot
7.4.1.5 Store
sample at 40
                                                        7.4.1.4
                                                        sample have
                                                   a high concentration
                                                    of formaldehyde?
                               7.4.1.4 Dilute
                             extract with mobile
                                  phase
  7.4.1.4 Exchange
solvent with methanol
    as in 7.3.5.5
7.4.1.4 Concentrate
  extract during
 solvent exchange
     process
                                        8315  - 33
                                   Revision  0
                             September  1994

-------
                 METHOD  8315
                  continued
     O
    7.4.2.2 - 7.4.2.3
Connect sample cartridge
  to dean syringe and
  place in syringe tack
        1
   7.4.2.4 Backflush
     carfrldgewith
      acetonitrite
^7.4.2,4 N.
Doesekiate N. Yes __
ttowb
bkx
.
No
eoome /
tod? /
7.4.24 Displace
trapped air will
it ni*tit*mtltmUit !•>
acetannmein
syringe using a long-tip
disposable Pasteur pjpet


   7.4.2.5 DflutB to 5
  mLwtthacetonitrile;
   label flask; pipat 2
     aNquoteinto
     sample viato
         I
     7.4.2.6 Store
     sample at 4C
                   8315  - 34
     Revision 0
September  1994

-------
                                      METHOD 8315
                                       continued
7.5.2 Set LC conditions
to produce appropriate
   retention times
        I
       7.5.1 Option
        1W2LC
       conditions?
    7.5.1 Set LC
conditions to produce
appropriate retention
   7.5.2.1 Filter and
 degas mobile phase
  7.6.2 Process calibration
  standards through same
processing steps as samples
                                    7.6.3 - 7.6.4
                                Analyze solvent blank
                              and calibration standards:
                                 tabulate peak areas
                               7.6.5 Determine response
                              factor at each concentration
                                      7.6.5
                                      Does
                                    calibration
                                   check meet
                                     criteria?
                                       7.6.5 Recalibrate
                                         Yes
                                                          Yes
                                            7.6.5
                                           Does
                                          __ IJik^MiJMM
                                          caiioraoon
                                         check meet
                                          criteria?
                                    0
                                      7.6.5 Prepare new
                                         calibration
                                         standards
                                       8315  -  35
                                                        Revision 0
                                                   September 1994

-------
                   METHOD 8315
                     continued
                                 O
1
f
7.6.6 - 7.6.7 Verify
calibration curve every day;
reanalyze 1 calibration
standard after 10
sample runs or less
                               7.7 Analyze samples
                                   byHPLC
 7.7.2 Inject a smaller
volume or dilute sample
     7.7.4 Further
  cleanup Is required
   7.7.2
 Does peak
area exceed
 calibration
  curve?
 7.7.4 Are
interferences
 present?
781 Calculate each
response factor, mean
response factor, and
percent BSD
i
i
7.8.2 - 7.8.5
Calculate analyte
concentrations
i

                                     Stop
                      831S -  36
                                 Revision  0
                           September  1994

-------
                                  APPENDIX A

            RECRYSTALLIZATION OF 2,4-DINITROPHENYLHYDRAZINE (DNPH)


NOTE: This  procedure should  be  performed under  a properly  ventilated hood.
      Inhalation of acetonitrile can result in nose and throat  irritation (brief
      exposure  at  500 ppm)  or more  serious  effects  at  higher concentration
      and/or longer exposures.

      A.I   Prepare a saturated solution of DNPH by boiling excess DNPH in 200 ml
of acetonitrile for approximately 1 hour.

      A.2   After 1 hour, remove and transfer the supernatant to a covered beaker
on  a  hot  plate and  allow  gradual  cooling  to 40  to  60°C.    Maintain  this
temperature range until 95% of the solvent has evaporated, leaving crystals.

      A.3   Decant the solution to waste and rinse the  remaining  crystals twice
with three times their apparent volume of acetonitrile.

      A.4   Transfer the crystals to a clean beaker, add 200 ml of acetonitrile,
heat to boiling, and again let the crystals grow slowly at 40  to 60°C until 95%
of the solvent has evaporated.  Repeat the rinsing process as in Sec. A.3.

      A.5   Take  an  aliquot  of  the  second   rinse,   dilute  10  times  with
acetonitrile, acidify  with  1 ml of 3.8  M perchloric acid per  100 mL of DNPH
solution, and analyze  with  HPLC as in  Sec.  7.0 for Option 2.   An acceptable
impurity level is less than 0.025 ng/jiL of formaldehyde in recrystallized DNPH
reagent or below the sensitivity (ppb, v/v) level indicated in Table 3 for the
anticipated sample volume.

      A.6   If the  impurity  level  is not satisfactory, pipet off  the solution to
waste, repeat the  recrystallization  as in Sec.  A.4  but rinse with  two  25 ml
portions of acetonitrile.  Prep and analyze the second rinse as  in Sec. A.5.

      A.7   When the impurity level j_s  satisfactory,  place  the  crystals in an
all-glass reagent bottle, add another  25 ml of  acetonitrile, stopper, and shake
the bottle.   Use clean pipets when removing the  saturated DNPH  stock solution to
reduce the possibility of contamination of the solution.  Maintain only a minimum
volume of the saturated solution adequate for day to day operation to minimize
waste of the purified reagent.
                                  8315  - 37                         Revision 0
                                                                September 1994

-------
                                  METHOD  8316

          ACRYLAMIDE,  ACRYLONITRILE AND ACROLEIN  BY HIGH  PERFORMANCE
                         LIQUID CHROMATOGRAPHY (HPLC)
1.0   SCOPE AND APPLICATION

      1.1   The following compounds can be determined by this method:



      Compound Name                                   CAS No.a


      Acrylamide                                       79-06-1
      Acrylonitrile                                   107-13-1
      Acrolein (Propenal)                             107-02-8


      a      Chemical Abstract Services Registry Number.

      1.2   The  method  detection  limits  (MDLs)  for  the  target  analytes  in
organic-free reagent water are listed in Table  1.  The method may be applicable
to other matrices.

      1.3   This method  is  restricted to use  by  or under the  supervision  of
analysts experienced in the  use  of high performance liquid chromatographs and
skilled in the interpretation of high performance  liquid chromatograms.   Each
analyst must demonstrate  the  ability  to generate  acceptable  results with this
method.
2.0   SUMMARY OF METHOD

      2.1   Water samples are analyzed by high performance liquid chromatography
(HPLC).  A  200  juL aliquot  is  injected  onto a C-18  reverse-phase  column,  and
compounds in the effluent are detected with an ultraviolet (UV) detector.


3.0   INTERFERENCES

      3.1   Contamination by carryover can occur whenever high-concentration and
low-concentration samples are sequentially analyzed.   To reduce carryover,  the
sample syringe must  be  rinsed  out between samples with solvent.   Whenever an
unusually concentrated  sample  is  encountered,  it should  be followed  by  the
analysis of solvent to check for cross contamination.
                                   8316 - 1                         Revision 0
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS
      4.1   HPLC system
            4.1.1 One  high  pressure pump.
            4.1.2 Octadecyl  Si lane  (ODS,  C-18)  reverse  phase HPLC  column,
      25 cm x 4.6 mm, 10 /xm, (Zorbax, or equivalent).
            4.1.3 Variable  wavelength UV detector.
            4.1.4 Data  system.
      4.2   Other apparatus
            4.2.1 Water degassing  unit  - 1 liter filter flask with stopper and
      pressure tubing.
            4.2.2 Analytical balance -  + 0.0001 g.
            4.2.3 Magnetic  stirrer and  magnetic stirring  bar.
            4.2.4 Sample filtration unit  -  syringe  filter with 0.45 /xm filter
      membrane, or equivalent disposable filter unit.
      4.3   Materials
            4.3.1 Syringes  - 10, 25, 50 and  250 /iL  and 10 mL.
            4.3.2 Volumetric pipettes,  Class A, glass  -1,5 and  10  mL.
            4.3.3 Volumetric flasks - 5,  10, 50 and  100 mL.
            4.3.4 Vials -   25  mL,  glass with Teflon lined screw caps or crimp
      tops.
5.0   REAGENTS
      5.1   Reagent grade inorganic chemicals shall be used in all tests. Unless
otherwise  indicated,  it  is  intended  that all  reagents  shall   conform  to the
specifications of the Committee on  Analytical Reagents  of  the American Chemical
Society, where  such specifications are available.  Other  grades may be used,
provided it is first ascertained that the reagent is  of  sufficiently high purity
to permit  its use without lessening the accuracy of the determination.
      5.2   Acrylamide, CH2:CHCONH2, 99+% purity, electrophoresis reagent grade.
      5.3   Acrylonitrile,  H2C:CHCN,  99+% purity.
      5.4   Acrolein, CH2:CHCHO, 99+7. purity.
                                   8316  - 2                         Revision 0
                                                                September  1994

-------
      5.5   Organic-free reagent water.   All  references to water in this method
refer to organic-free reagent water,  as defined  in Chapter One.  Sparge with He
to eliminate 02 to prevent significant absorption interference from 02  at the 195
nm wavelength.

      5.6   Stock  standard  solutions  -  Can  be prepared  from pure standard
materials or  can  be  purchased as certified  solutions.   Commercially prepared
stock standards  can  be  used if  they are certified  by  the  manufacturer and
verified against a standard made from pure material.

            5.6.1  Aery1 amide

                   5.6.1.1     Weigh 0.0100 g  of acrylamide neat standard into a
            100 ml volumetric flask,  and dilute to  the mark with  organic-free
            reagent water. Calculate  the  concentration of the standard solution
            from the actual weight used.  When compound purity  is assayed to be
            96%  or greater,  the weight  can be  used without correction  to
            calculate the concentration of the stock standard.

                   5.6.1.2     Transfer the stock solution  into vials with Teflon
            lined screw caps or  crimp tops.  Store at 4°C, protected from light.

                   5.6.1.3     Stock solutions must be replaced after six months,
            or  sooner  if  comparison with   the  check standards  indicates  a
            problem.

            5.6.2  Acrylonitrile and Acrolein - Prepare separate stock solutions
      for acrylonitrile and  acrolein.

                   5.6.2.1     Place about  9.8 ml of organic-free reagent water
            into a 10 ml  volumetric flask before  weighing  the flask and stopper.
            Weigh the flask and  record the weight to the nearest 0.0001 g.  Add
            two drops of  neat standard,  using  a 50 juL syringe, to  the flask.
            The liquid must fall directly into the water, without contacting the
            inside wall  of the flask.

                   CAUTION:    Acrylonitrile  and  acrolein  are  toxic.   Standard
                              preparation  should be  performed in an  laboratory
                              fume hood.

                   5.6.2.2     Stopper the flask and  then reweigh.   Dilute to
            volume with organic-free  reagent water.   Calculate the concentration
            from the  net  gain in weight.  When compound purity  is assayed to be
            96%  or greater,  the weight  can be  used without correction  to
            calculate the concentration of the stock standard.

                   5.6.2.3     Stock solutions must be replaced after six months,
            or  sooner  if  comparison with   the  check standards  indicates  a
            problem.
                                   8316 - 3                         Revision 0
                                                                September 1994

-------
      5.7   Calibration standards

            5.7.1 Prepare   calibration  standards   at   a  minimum   of  five
      concentrations by diluting the stock solutions with organic-free reagent
      water.

            5.7.2 One calibration standard should be prepared at a concentration
near, but  above,  the method  detection limit;  the  remaining  standards  should
correspond to the range of  concentrations  found in real samples, but should not
exceed the working range of the HPLC system (1  mg/L to 10 mg/L).


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this  chapter,  Organic  Analytes,
Sec. 4.1.
7.0
8.0
PROCEDURE

7.1   HPLC Conditions

Mobile Phase:
Injection Volume:
Flow Rate:
Pressure:
Temperature:
Detector UV wavelength:

7.2   Calibration:
                                    Degassed organic-free reagent water
                                    200 ML
                                    2.0 mL/min
                                    38 atm
                                    25°C
                                    195 nm
            7.2.1  Prepare standard solutions of acrylamide as described in Sec.
      5.7.1.  Inject 200 ,uL  aliquots  of each solution into the chromatograph.
      See Method 8000  for  additional  guidance on calibration  by  the external
      standard method.

      7.3   Chromatographic analysis:

            7.3.1  Analyze the samples  using the same chromatographic conditions
      used to prepare the standard  curve.  Suggested chromatographic conditions
      are given  in  Sec.  7.1.   Table 1  provides the retention  times  that were
      obtained under these conditions  during method development.
QUALITY CONTROL

8.1   Refer to Chapter One for specific quality control procedures.
      8.2   Before processing any samples, the analyst must demonstrate, through
the analysis of a method blank, that all glassware and reagents are interference
free.
                                   8316 - 4
                                                              Revision 0
                                                          September 1994

-------
9.0   METHOD PERFORMANCE

      9.1   Method performance data are not available.


10.0  REFERENCES

1.    Hayes, Sam;  "Aery!amide, Acrylonitrile,  and Acrolein  Determination in
      Water by High Pressure Liquid Chromatography," USEPA.
                                   8316 - 5                         Revision 0
                                                                September 1994

-------
                                    TABLE  1
             ANALYTE RETENTION TIMES AND METHOD DETECTION LIMITS
                                  Retention             MDL
Compound                          Time (min)           (M9/L)
Acrylamide                           3.5                10
Acrylonitrile                        8.9                20
Acrolein (Propenal)                 10.1                30
                                    8316  -  6                          Revision  0
                                                                 September  1994

-------
                          METHOD 8316
ACRYLAMIDE. ACRYLONITRILE  AND ACROLEIN BY HIGH  PERFORMANCE
                 LIQUID CHRQMATOGRAPHY  (HPLC)
                   (   Start   J
                      7.1 Set by
                        HPLC
                      Conditions.
                     7.2 Calibrate
                    Chromatograph.
                         7.3
                   Chromatographic
                      analysis.
                    (   Stop   J
                             8316 - 7
    Revision 0
September 1994

-------
                                  METHOD  8318

                N-METHYLCARBAMATES BY HIGH PERFORMANCE LIQUID
                             CHROMATOGRAPHY  (HPLC)
1.0   SCOPE AND APPLICATION

      1.1   Method   8318   is   used   to   determine   the  concentration   of
N-methylcarbamates in soil, water and waste matrices. The following compounds can
be determined by this method:
      Compound Name                                               CAS No.a


      Aldicarb (Temik)                                              116-06-3
      Aldicarb Sulfone                                             1646-88-4
      Carbaryl (Sevin)                                               63-25-2
      Carbofuran (Furadan)                                         1563-66-2
      Dioxacarb                                                    6988-21-2
      3-Hydroxycarbofuran                                         16655-82-6
      Methiocarb (Mesurol)                                         2032-65-7
      Methomyl (Lannate)                                          16752-77-5
      Promecarb                                                    2631-37-0
      Propoxur (Baygon)                                             114-26-1


      a  Chemical Abstract Services Registry Number.

      1.2   The method detection limits (MDLs) of Method 8318 for determining the
target analytes in organic-free reagent water and in soil are listed  in Table  1.

      1.3   This method is restricted to  use by,  or under the supervision of,
analysts experienced in the use of high performance liquid chromatography (HPLC)
and  skilled  in  the   interpretation  of  chromatograms.     Each  analyst  must
demonstrate the ability to generate acceptable  results with this method.


2.0   SUMMARY OF METHOD

      2.1   N-methylcarbamates are extracted from aqueous samples with methylene
chloride, and  from  soils,  oily solid waste and  oils with  acetonitrile.   The
extract solvent is exchanged to methanol/ethylene glycol,  and then the extract
is cleaned  up  on  a  C-18 cartridge, filtered,  and eluted  on a C-18 analytical
column.  After separation, the target  analytes  are  hydrolyzed and derivatized
post-column, then quantitated fluorometrically.

      2.2   Due to  the specific  nature  of this  analysis,  confirmation  by a
secondary method  is  not essential.  However,  fluorescence  due  to  post-column
derivatization may be  confirmed by substituting the NaOH  and o-phthalaldehyde
solutions with organic-free  reagent  water and  reanalyzing  the  sample.    If


                                   8318 - 1                         Revision 0
                                                                September  1994

-------
fluorescence is still  detected, then a positive interference is present and care
should be taken in the interpretation of the results.

      2.3   The  sensitivity of  the  method  usually depends on  the  level  of
interferences present, rather than on the instrumental conditions.  Waste samples
with a high  level  of extractable fluorescent compounds  are  expected to yield
significantly higher detection limits.


3.0   INTERFERENCES

      3.1   Fluorescent compounds, primarily  alkyl  amines and  compounds which
yield  primary  alkyl  amines  on   base hydrolysis,  are  potential   sources  of
interferences.

      3.2   Coeluting compounds  that  are  fluorescence  quenchers  may result in
negative interferences.

      3.3   Impurities  in  solvents  and  reagents   are  additional   sources  of
interferences.   Before processing  any  samples,  the analyst must  demonstrate
daily, through the analysis  of solvent blanks, that  the  entire analytical system
is interference free.
4.0   APPARATUS AND MATERIALS

      4.1   HPLC system

            4.1.1 An  HPLC  system  capable  of  injecting  20  juL  aliquots  and
      performing multilinear gradients at  a constant flow.  The system must also
      be equipped with a data system to measure the peak areas.

            4.1.2 C-18 reverse phase HPLC column, 25 cm x 4.6 mm  (5 /zm).

            4.1.3 Post Column Reactor with two solvent delivery  systems (Kratos
      PCRS  520  with two  Kratos  Spectroflow 400 Solvent Delivery Systems,  or
      equivalent).

            4.1.4 Fluorescence detector (Kratos Spectroflow 980, or equivalent).

      4.2   Other apparatus

            4.2.1 Centrifuge.

            4.2.2 Analytical balance  - +  0.0001 g.

            4.2.3 Top loading balance  - + 0.01 g.

            4.2.4 Platform  shaker.

            4.2.5 Heating block, or equivalent apparatus, that  can accommodate
      10 mL graduated vials  (Sec. 4.3.11).
                                   8318  - 2                         Revision 0
                                                                September  1994

-------
      4.3   Materials

            4.3.1 HPLC injection syringe - 50 ,uL.

            4.3.2 Filter paper,  (Whatman #113 or #114, or equivalent).

            4.3.3 Volumetric pipettes, Class A, glass,  assorted sizes.
                                                         p
            4.3.4 Reverse phase cartridges,  (C-18 Sep-Pak [Waters Associates],
      or equivalent).

            4.3.5 Glass syringes - 5 ml.

            4.3.6 Volumetric flasks, Class A - Sizes as appropriate.

            4.3.7 Erlenmeyer flasks with teflon-lined screw caps, 250 ml.

            4.3.8 Assorted glass funnels.

            4.3.9 Separatory funnels,  with ground  glass stoppers  and teflon
      stopcocks - 250 ml.

            4.3.10      Graduated cylinders - 100 ml.

            4.3.11      Graduated glass vials - 10 ml,  20 ml_.

            4.3.12      Centrifuge tubes - 250 ml.

            4.3.13      Vials  -  25 ml,  glass  with  Teflon  lined screw caps or
      crimp tops.

            4.3.14      Positive  displacement  micro-pipettor,  3  to  25  /zL
      displacement, (Gilson Microman [Rainin #M-25] with  tips,  [Rainin #CP-25],
      or equivalent).

            4.3.15      Nylon  filter unit,  25 mm  diameter,  0.45 /zm pore size,
      disposable (Alltech Associates,  #2047, or equivalent).


5.0   REAGENTS

      5.1   HPLC grade chemicals shall be used in  all tests.  It is  intended that
all reagents shall  conform to the specifications of the Committee on Analytical
Reagents  of the  American  Chemical  Society,  where such   specifications  are
available.  Other grades may be used,  provided  it  is first ascertained that the
reagent is of sufficiently  high  purity  to  permit  its use without lowering the
accuracy of the determination.

      5.2   General

            5.2.1 Acetonitrile, CH,CN  -  HPLC grade -  minimum UV cutoff at 203 nm
      (EM Omnisolv #AX0142-1, or equivalent).
                                   8318 - 3                         Revision 0
                                                                September 1994

-------
      5.2.2 Methanol, CH3OH - HPLC grade - minimum UV cutoff at 230 nm (EM
Omni solv #MX0488-1, or equivalent).

      5.2.3 Methylene chloride,  CH?C1?  - HPLC grade - minimum UV cutoff at
230 nm (EM Omnisolv #0X0831-1,  or Equivalent).

      5.2.4 Hexane, CgH,, - pesticide grade - (EM Omnisolv #HX0298-1, or
equivalent).

      5.2.5 Ethylene glycol,  HOCH2CH2OH -  Reagent grade -  (EM Science, or
equivalent).

      5.2.6 Organic-free reagent water - All  references to water in this
method refer to organic-free reagent water, as defined in Chapter One.

      5.2.7 Sodium hydroxide, NaOH -  reagent  grade - 0.05N NaOH solution.

      5.2.8 Phosphoric acid,  H3P04 - reagent grade.

      5.2.9 pH 10 borate buffer (J.T. Baker #5609-1, or equivalent).

      5.2.10      o-Phthalaldehyde,  o-CcMCHO),, - reagent grade (Fisher
#0-4241, or equivalent).               ° H     c

      5.2.11      2-Mercaptoethanol, HSCH2CH2OH - reagent grade (Fisher
#0-3446, or equivalent).

      5.2.12      N-methylcarbamate  neat  standards   (equivalence  to EPA
standards must be demonstrated  for purchased solutions).

      5.2.13      Chloroacetic  acid, C1CH2COOH, 0.1  N.

5.3   Reaction solution

      5.3.1 Dissolve 0.500 g of  o-phthalaldehyde in  10 ml  of methanol, in
a  1  L  volumetric flask.   To this solution,  add 900  ml  of organic-free
reagent water,  followed  by 50  mL of the  borate buffer  (pH  10).   After
mixing well, add  1  ml  of 2-mercaptoethanol,  and dilute to the mark with
organic-free reagent water.  Mix the solution thoroughly.  Prepare fresh
solutions  on  a weekly basis, as  needed.   Protect  from  light  and store
under refrigeration.

5.4   Standard solutions

      5.4.1 Stock   standard  solutions:  prepare  individual   1000  mg/L
solutions by adding 0.025 g of carbamate to a 25 mL  volumetric flask, and
diluting to the mark with methanol.  Store  solutions,  under refrigeration,
in glass vials with Teflon lined screw caps or crimp tops.  Replace every
six months.

      5.4.2 Intermediate  standard solution:   prepare a mixed  50.0 mg/L
solution by adding  2.5 mL of each stock  solution to a 50 mL volumetric
flask,  and diluting to  the  mark with methanol.  Store  solutions, under


                             8318 -  4                         Revision 0
                                                          September  1994

-------
      refrigeration, in glass vials with  Teflon  lined screw caps or crimp tops.
      Replace every three months.

            5.4.3 Working standard solutions: prepare 0.5, 1.0,  2.0, 3.0 and 5.0
      mg/L  solutions  by  adding  0.25,  0.5,  1.0,  1.5  and   2.5  mL  of  the
      intermediate mixed  standard to respective 25 ml  volumetric  flasks,  and
      diluting  each  to  the  mark  with  methanol.   Store  solutions,  under
      refrigeration, in glass vials with  Teflon  lined screw caps or crimp tops.
      Replace every two months, or sooner if necessary.

            5.4.4 Mixed QC standard solution: prepare a  40.0 mg/L solution from
      another  set  of  stock  standard solutions,  prepared similarly  to  those
      described in Sec.  5.4.1.   Add 2.0  ml of each stock solution to  a  50 ml
      volumetric  flask  and  dilute  to  the  mark  with  methanol.   Store  the
      solution, under refrigeration, in a glass vial with a Teflon lined screw
      cap or crimp top.  Replace every three months.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Due to  the  extreme instability of  N-methylcarbamates  in  alkaline
media, water, waste water  and  leachates  should  be  preserved  immediately after
collection by acidifying to pH 4-5 with 0.1 N chloroacetic acid.

      6.2   Store samples at 4°C  and  out of  direct sunlight,  from  the time of
collection  through  analysis.   N-methylcarbamates  are  sensitive to  alkaline
hydrolysis and heat.

      6.3   All samples must be extracted within seven days of collection,  and
analyzed within 40 days of extraction.


7.0   PROCEDURE

      7.1   Extraction

            7.1.1 Water, domestic  wastewater,  aqueous  industrial wastes,  and
      leachates

                  7.1.1.1     Measure 100 mL  of  sample into a 250 mL separatory
            funnel  and extract  by  shaking vigorously for  about 2 minutes with 30
            mL of methylene  chloride.   Repeat  the  extraction  two  more  times.
            Combine all three extracts in a 100 mL volumetric  flask and dilute
            to volume with methylene chloride.   If cleanup  is  required,  go to
            Sec.  7.2.   If cleanup  is not  required, proceed directly  to Sec.
            7.3.1.

            7.1.2 Soils, solids, sludges, and heavy aqueous suspensions

                  7.1.2.1     Determination of  sample %  dry weight - In certain
            cases,  sample results are desired based on dry-weight basis.   When
            such  data  is  desired,  a portion  of sample  for this determination
            should be  weighed  out  at the  same time  as the  portion  used  for
            analytical determination.

                                   8318 - 5                         Revision 0
                                                                September 1994

-------
            WARNING:    The drying oven should be contained in a hood or
                        vented.  Significant laboratory contamination may
                        result  from  a heavily  contaminated  hazardous
                        waste sample.

                  7.1.2.1.1   Immediately after weighing  the  sample for
            extraction, weigh 5-10 g of the sample into a tared crucible.
            Determine the % dry weight of the sample by drying overnight
            at 105°C.   Allow to cool  in a  desiccator before weighing:

                  % dry weight = q of dry sample x 100
                                   g of sample

            7.1.2.2     Extraction - Weigh out 20 + 0.1 g of sample into
      a 250 ml Erlenmeyer flask with a Teflon-lined  screw cap.  Add 50 ml
      of acetonitrile and shake for 2 hours on a platform shaker.  Allow
      the mixture to settle  (5-10 min), then decant the extract into a 250
      ml centrifuge tube.  Repeat the extraction two more times with 20 ml
      of acetonitrile and 1  hour  shaking  each  time.   Decant and combine
      all three extracts. Centrifuge the  combined extract at 200 rpm for
      10 min.  Carefully decant the supernatant into a 100 mL volumetric
      flask and dilute  to volume with acetonitrile.  (Dilution factor = 5)
      Proceed to Sec. 7.3.2.

      7.1.3 Soils heavily contaminated with non-aqueous substances, such
as oils

            7.1.3.1     Determination  of  sample %  dry weight  - Follow
      Sees. 7.1.2.1 through 7.1.2.1.1.

            7.1.3.2     Extraction - Weigh out 20 + 0.1 g of sample into
      a 250 ml Erlenmeyer flask with a Teflon-lined  screw cap.  Add 60 ml
      of hexane and shake for  1  hour  on  a platform shaker. Add  50 ml of
      acetonitrile and  shake for  an additional  3 hours. Allow the mixture
      to settle (5-10 min), then decant the solvent layers into a 250 mL
      separatory funnel.  Drain  the  acetonitrile  (bottom layer) through
      filter paper into a 100 ml volumetric flask.  Add 60 mL of hexane and
      50 mL of acetonitrile to the sample extraction flask and shake for
      1 hour.  Allow the mixture to settle, then decant the mixture into
      the  separatory  funnel  containing  the  hexane  from  the  first
      extraction.   Shake the  separatory  funnel  for 2 minutes,  allow the
      phases  to  separate,  drain  the  acetonitrile  layer  through filter
      paper  into  the   volumetric  flask,  and  dilute  to  volume  with
      acetonitrile.  (Dilution factor = 5)  Proceed to Sec. 7.3.2.

      7.1.4 Non-aqueous liquids such as oils

            7.1.4.1     Extraction - Weigh out 20 + 0.1 g of sample into
      a  125  rnL separatory  funnel.   Add  40  mL of  hexane and  25  mL of
      acetonitrile and vigorously shake the sample mixture  for 2 minutes.
      Allow the phases  to  separate,  then  drain the acetonitrile (bottom
      layer) into a 100 mL  volumetric flask. Add 25  mL of acetonitrile to
      the  sample  funnel,  shake  for  2  minutes,  allow  the  phases  to


                             8318 - 6                         Revision 0
                                                          September 1994

-------
            Repeat the extraction with  another  25  ml  portion of acetonitrile,
            combining  the  extracts.    Dilute  to   volume  with  acetonitrile.
            (Dilution factor = 5).  Proceed to Sec. 7.3.2.

      7.2   Cleanup  -  Pipet 20.0 ml  of the extract  into  a 20 ml  glass vial
containing 100 /uL of ethylene glycol.   Place the vial  in a heating block  set at
50° C, and gently evaporate  the  extract under a stream of  nitrogen (in a fume
hood) until  only the ethylene glycol  keeper remains.  Dissolve  the ethylene
glycol residue in 2 ml of methanol,  pass the extract through a pre-washed C-18
reverse phase  cartridge,  and collect the  eluate in a 5 ml  volumetric  flask.
Elute the cartridge with methanol, and collect the eluate until the final volume
of 5.0 mL  is  obtained.   (Dilution factor = 0.25)   Using a  disposable 0.45 /urn
filter,  filter an aliquot  of  the clean extract directly into a properly labelled
autosampler vial.  The extract is now  ready for  analysis.  Proceed to Sec. 7.4.

      7.3   Solvent Exchange

            7.3.1 Water,   domestic wastewater,  aqueous industrial  wastes,  and
      leachates:

            Pipet 10.0 ml of the extract  into  a  10  ml graduated  glass vial
      containing 100 /uL of ethylene glycol.  Place the vial in a heating block
      set at 50  C,  and gently evaporate the  extract under a stream of nitrogen
      (in  a  fume hood) until only  the  ethylene glycol  keeper remains.   Add
      methanol to the ethylene glycol  residue, dropwise, until the total volume
      is 1.0 ml.  (Dilution factor = 0.1).   Using a disposable 0.45 jum filter,
      filter this extract directly  into a  properly labelled autosampler vial.
      The extract is now ready for analysis.  Proceed to Sec. 7.4.

            7.3.2 Soils,   solids,  sludges,  heavy aqueous  suspensions,  and non-
      aqueous liquids:

            Elute 15 ml of the acetonitrile extract  through  a C-18 reverse phase
      cartridge, prewashed with  5 ml of  acetonitrile.  Discard the  first 2 ml of
      eluate and collect the  remainder.  Pipet 10.0  ml  of the clean extract into
      a 10 ml graduated glass vial containing 100 juL of ethylene glycol.  Place
      the vial in a heating block set at 50° C, and  gently evaporate the extract
      under a stream of nitrogen  (in a fume  hood) until  only the ethylene glycol
      keeper remains.  Add methanol to  the ethylene glycol  residue, dropwise,
      until the  total  volume is  1.0  ml.   (Additional dilution  factor = 0.1;
      overall dilution factor = 0.5).  Using a disposable 0.45 jum filter, filter
      this extract  directly into a properly labelled autosampler  vial.   The
      extract is now ready for analysis.  Proceed to Sec.  7.4.

      7.4   Sample Analysis

            7.4.1 Analyze the  samples  using  the  chromatographic  conditions,
      post-column reaction parameters and instrument parameters given in Sees.
      7.4.1.1, 7.4.1.2, 7.4.1.3  and 7.4.1.4.    Table  2 provides  the retention
      times that were obtained under these conditions during method development.
      A chromatogram of the separation  is shown  in  Figure 1.
                                   8318 - 7                         Revision 0
                                                                September 1994

-------
            7.4.1.1
      Chromatographic  Conditions  (Recommended)
            Solvent A:         Organic-free  reagent water,  acidified with
                              0.4  mL  of phosphoric  acid  per  liter of
                              water
            Solvent B:         Methanol/acetonitrile  (1:1,  v/v)
            Flow rate:         1.0  mL/min
            Injection Volume:  20 juL
            Solvent delivery system  program:
      36.02
Function
   FR
   B%
   B%
   B%
   B%
   B%
   B%
 ALARM
                                          Duration
                                          (min)
20
 5
 5
 3
 7
 0.01
            7.4.1.2
      Post-column Hydrolysis  Parameters  (Recommended)
            Solution:
            Flow Rate:
            Temperature:
            Residence  Time:
            0.05 N aqueous  sodium  hydroxide
            0.7 mL/min
            95° C
            7.4.1.3
      (Recommended)

            Solution:
            35 seconds (1  mL reaction  coil)

      Post-column    Derivatization    Parameters
            Flow Rate:
            Temperature:
            Residence time:
            o-phthalaldehyde/2-mercaptoethanol
            5.3.1)
            0.7 mL/min
            40° C
            25 seconds (1  mL reaction  coil)
                       (Sec.
            7.4.1.4
      Fluorometer Parameters  (Recommended)
            Cell:                    10 ML
            Excitation wavelength:   340 nm
            Emission wavelength:     418 nm cutoff filter
            Sensitivity wavelength:  0.5 juA
            PMT voltage:            -800 V
            Time constant:           2 sec

      7.4.2 If the peak areas of the  sample signals exceed the calibration
range of the  system,  dilute  the  extract as necessary and reanalyze  the
diluted extract.
                             8318 - 8
                                            Revision  0
                                        September 1994

-------
7.5   Calibration:

      7.5.1 Analyze a solvent  blank (20 juL of methanol)  to ensure that the
system is clean.  Analyze the calibration standards  (Sec. 5.4.3), starting
with the 0.5 mg/L standards and ending with the 5.0  mg/L standard.  If the
percent  relative  standard  deviation  (%RSD) of the  mean response factor
(RF) for each  analyte  does not exceed  20%,  the system  is  calibrated and
the analysis of samples may proceed.   If the %RSD for any analyte exceeds
20%,  recheck  the  system  and/or  recalibrate  with  freshly  prepared
calibration solutions.

      7.5.2 Using the established calibration mean response factors, check
the  calibration  of  the  instrument  at  the  beginning of  each  day  by
analyzing  the  2.0 mg/L mixed standard.   If  the concentration  of  each
analyte falls within  the range of 1.70 to 2.30 mg/L  (i.e., within + 15% of
the true value),  the instrument  is considered to be calibrated  and the
analysis of  samples  may proceed.   If the observed  value of any  analyte
exceeds  its true  value by  more  than  ±  15%,  the  instrument must  be
recalibrated (Sec. 7.5.1).

      7.5.3 After 10 sample runs,  or  less,  the  2.0 mg/L  standards must be
analyzed to ensure that the retention times and response factors are still
within  acceptable  limits.    Significant  variations   (i.e.,  observed
concentrations exceeding the true concentrations by more than + 15%) may
require a re-analysis of the samples.

7.6   Calculations

      7.6.1 Calculate each response factor as follows (mean value based on
5 points):

            concentration of standard
      RF =  	
              area of the signal

                       5
                       (I RF.)

      mean RF = RF = 	
                   [(I RFi - RF)2]1/2 / 4
              	     i
      %RSD of RF = 	—	     X 100%
                            RF

      7.6.2 Calculate  the concentration  of  each  N-methylcarbamate  as
follows:

      jug/g or mg/L =  (RF) (area of signal)  (dilution factor)
                             8318 - 9                         Revision 0
                                                          September 1994

-------
8.0   QUALITY CONTROL

      8.1   Before processing any samples, the analyst must demonstrate, through
the analysis  of  a method blank for  each  matrix type, that  all  glassware and
reagents are  interference free.    Each time there is  a  change  of reagents,  a
method blank must be processed as a safeguard against laboratory contamination.

      8.2   A QC check solution must be prepared and analyzed with each sample
batch that is processed.   Prepare this  solution, at a concentration of 2.0 mg/L
of each analyte,  from the 40.0 mg/L mixed QC standard solution (Sec. 5.4.4).  The
acceptable response range is 1.7 to 2.3 mg/L for each analyte.

      8.3   Negative interference due to quenching may be examined by spiking the
extract with  the appropriate  standard,  at  an  appropriate  concentration, and
examining the observed response against the expected response.

      8.4   Confirm  any  detected analytes  by  substituting  the NaOH  and OPA
reagents in the post column  reaction  system with deionized water, and reanalyze
the suspected  extract.   Continued  fluorescence response will  indicate that  a
positive interference is present (since the fluorescence  response  is not due to
the post column derivatization). Exercise caution  in the interpretation of the
chromatogram.


9.0   METHOD PERFORMANCE

      9.1   Table 1 lists the single operator method detection limit (MDL) for
each  compound  in organic-free  reagent water  and  soil.    Seven/ten  replicate
samples were  analyzed,  as indicated in the  table.   See reference  7  for more
details.

      9.2   Tables  2, 3  and  4 list  the single operator average recoveries and
standard deviations for  organic-free reagent  water,  wastewater  and  soil.   Ten
replicate samples were analyzed at each indicated spike concentration for each
matrix type.

      9.3   The method detection limit, accuracy and  precision obtained will be
determined by the sample matrix.


10.0  REFERENCES

1.    California Department of Health Services,  Hazardous Materials Laboratory,
      "N-Methylcarbamates by HPLC", Revision No. 1.0, September 14,  1989.

2.    Krause, R.T.  Journal of Chromatographic Science, 1978, vol. 16, pg  281.

3.    Klotter,  Kevin,  and   Robert  Cunico,  "HPLC  Post Column  Detection  of
      Carbamate  Pesticides", Varian Instrument Group, Walnut Creek,  CA  94598.

4.    USEPA,  "Method  531.     Measurement of  N-Methylcarbomyloximes  and  N-
      Methylcarbamates in Drinking Water  by Direct Aqueous  Injection HPLC with
      Post Column Derivatization",   EPA 600/4-85-054, Environmental  Monitoring
      and Support Laboratory, Cincinnati, OH  45268.

                                   8318 -  10                         Revision 0
                                                                September 1994

-------
5.    USEPA, "Method 632.  The Determination of Carbamate and Urea Pesticides in
      Industrial  and Municipal  Wastewater",  EPA  600/4-21-014,  Environmental
      Monitoring and Support Laboratory, Cincinnati, OH  45268.

6.    Federal Register, "Appendix B to Part 136 - Definition and Procedure for
      the Determination of the Method  Detection  Limit - Revision 1.11", Friday,
      October 26, 1984, 49, No. 209, 198-199.

7.    Okamoto, H.S., D.  Wijekoon, C. Esperanza, J. Cheng,  S. Park, J. Garcha, S.
      Gill,  K.  Perera "Analysis for  N-Methylcarbamate Pesticides  by  HPLC in
      Environmental Samples",  Proceedings of the Fifth Annual USEPA Symposium on
      Waste Testing and Quality Assurance,  July 24-28,  1989,  Vol. II, 57-71.
                                  8318  -  11                         Revision 0
                                                                September 1994

-------
                                   TABLE  1
                     ELUTION ORDER, RETENTION TIMESa AND
                    SINGLE OPERATOR METHOD DETECTION  LIMITS
Method Detection Limits
Compound


Aldicarb Sulfone
Methomyl (Lannate)
3-Hydroxycarbofuran
Dioxacarb
Aldicarb (Temik)
Propoxur (Baygon)
Carbofuran (Furadan)
Carbaryl (Sevin)
cr-Naphthor
Methiocarb (Mesurol)
Promecarb
Retention
Time
(min)
9.59
9.59
12.70
13.50
16.05
18.06
18.28
19.13
20.30
22.56
23.02
Organic-free
Reagent Water
(M9/L)
1.9C
1.7
2.6
2.2
9.4C
2.4
2.0
1.7
-
3.1
2.5

Soil
(M9/kg)
44C
12c
10c
>50C
12C
17
22
31
-
32
17
a

b



c

d
See Sec. 7.4 for chromatographic conditions

MDL  for  organic-free  reagent  water,  sand,   soil  were  determined  by
analyzing  10  low  concentration  spike  replicate  for  each matrix  type
(except where noted).  See reference 7 for more details.

MDL determined by analyzing 7 spiked replicates.

Breakdown product of Carbaryl.
                                   8318 - 12
                                                              Revision 0
                                                          September 1994

-------
                                TABLE 2
                  SINGLE OPERATOR AVERAGE RECOVERY AND
            PRECISION DATA3 FOR ORGANIC-FREE REAGENT WATER
Compound
Aldicarb Sulfone
Methomyl (Lannate)
3-Hydroxycarbofuran
Dioxacarb
Aldicarb (Temik)
Propoxur (Baygon)
Carbofuran (Furadan)
Carbaryl (Sevin)
Methiocarb (Mesurol)
Promecarb
Recovered
225
244
210
241
224
232
239
242
231
227
% Recovery
75.0
81.3
70.0
80.3
74.7
77.3
79.6
80.7
77.0
75.7
SD
7.28
8.34
7.85
8.53
13.5
10.6
9.23
8.56
8.09
9.43
%RSD
3.24
3.42
3.74
3.54
6.03
4.57
3.86
3.54
3.50
4.1
Spike Concentration = 300 jug/L of each compound, n = 10
                               8318 - 13                        Revision 0
                                                            September  1994

-------
                             TABLE 3
              SINGLE OPERATOR AVERAGE RECOVERY AND
                 PRECISION DATAa FOR WASTEWATER
Compound
Aldicarb Sulfone
Methomyl (Lannate)
3-Hydroxycarbofuran
Dioxacarb
Aldicarb (Temik)
Propoxur (Baygon)
Carbofuran (Furadan)
Carbaryl (Sevin)
Methiocarb (Mesurol)
Promecarb
a Spike Concentration
KI/% w«ns*s*t*«*\vttf
Recovered
235
247
2§J
D
258
263
262
262
254
263
= 300 jug/L of each

% Recovery
78.3
82.3
83.7
-
86.0
87.7
87.3
87.3
84.7
87.7
compound, n = 10

SD
17.6
29.9
25.4
-
16.4
16.7
15.7
17.2
19.9
15.1


%RSD
7.49
12.10
10.11
-
6.36
6.47
5.99
6.56
7.83
5.74


recovery
                            8318 - 14
    Revision 0
September 1994

-------
                                TABLE 4
                 SINGLE OPERATOR AVERAGE RECOVERY AND
                       PRECISION DATA3 FOR SOIL
Compound
Aldicarb Sulfone
Methomyl (Lannate)
3-Hydroxycarbofuran
Dioxacarb
Aldicarb (Temik)
Propoxur (Baygon)
Carbofuran (Furadan)
Carbaryl (Sevin)
Methiocarb (Mesurol)
Promecarb
Recovered
1.57
1.48
1.60
1.51
1.29
1.33
1.46
1.53
1.45
1.29
% Recovery
78.5
74.0
80.0
75.5
64.5
66.5
73.0
76.5
72.5
64.7
SD
0.069
0.086
0.071
0.073
0.142
0.126
0.092
0.076
0.071
0.124
%RSD
4.39
5.81
4.44
4.83
11.0
9.47
6.30
4.90
4.90
9.61
Spike Concentration = 2.00 mg/kg of each compound,  n = 10
                               8318 -  15
    Revision 0
September 1994

-------
                                   FIGURE 1
   100-
R
E
S
P
0
N
S
E
                              TIME  (MIN)
1.00 ]ug/mL EACH OF:

1.  ALDICARB  SULFONE

2.  METHOMYL

3.  3-HYDROXYCARBOFURAN

4.  DIOXACARB

5.  ALDICARB
6.   PROPOXUR

7.   CARBOFURAN

8.   CARBARYL

9.   METHIOCARB

10.  PROMECARB
                                  8318  -  16
                       Revision 0
                  September  1994

-------
   >-
   :n
   Q-
   CD
   O
   O
   *-H


   o-
00
•—i LU
co o
CO Z

o z:
o on
   <


   QQ


   O
   _J
   >-
   n:

   LU
   z:
     i
   z
                                                                                                                           C Ok
                                                                                                                           O —i

                                                                                                                           I/i s_
                                                                                                                           •I— Q)
                                                                                                                           > -Q
                                                                                                                           0) E
                                                                                                                           Q£ O)
 i

CO
f—I
CO
00

-------
                                                                                          C 01
                                                                                          O <-*
00—»
-H T3
CO 0)
00 3

O -r-
O *J
z c
I— O
uj u
PI
                                                                               8s
                                                                                          > JD
                                                                                          O) S
                                                                                          Q£ 0)
                                                                                            O)
                           00
oo

-------
                                  METHOD  8321

                SOLVENT EXTRACTABLE NON-VOLATILE COMPOUNDS BY
     HIGH PERFORMANCE LIQUID CHRQMATOGRAPHY/THERMOSPRAY/MASS SPECTROMETRY
                 (HPLC/TSP/MS) OR ULTRAVIOLET (UV) DETECTION
1.0   SCOPE AND APPLICATION

      1.1   This method covers the use of high performance liquid chromatography
(HPLC),  coupled with  either thermospray-mass  spectrometry  (TSP-MS),  and/or
ultraviolet (UV), for the determination of disperse azo dyes, organophosphorus
compounds, and  Tris-(2,3-dibromopropyl)phosphate  in  wastewater,  ground water,
sludge, and soil/sediment matrices,  and  chlorinated  phenoxyacid  compounds and
their esters in wastewater,  ground water,  and soil/sediment matrices.  Data are
also provided for chlorophenoxy  acid  herbicides  in fly ash (Table 15), however,
recoveries for most compounds are very poor indicating  poor extraction efficiency
for  these  analytes  using the  extraction procedure  included in this  method.
Additionally,  this method may  apply  to other non-volatile  compounds that are
solvent extractable,  are amenable to  HPLC, and are ionizable under thermospray
introduction for mass spectrometric detection.   The following compounds can be
determined by this method:
      Compound Name
      Fluorescent Brightener 61
      Fluorescent Brightener 236

      Alkaloids
      Caffeine
      Strychnine
  CAS No.'
Azo Dyes
Disperse Red 1
Disperse Red 5
Disperse Red 13
Disperse Yellow 5
Disperse Orange 3
Disperse Orange 30
Disperse Brown 1
Solvent Red 3
Solvent Red 23
Anthraquinone Dyes
Disperse Blue 3
Disperse Blue 14
Disperse Red 60
Coumarin Dyes
(Fluorescent Briqhteners)

2872-52-8
3180-81-2
2832-40-8
6439-53-8
730-40-5
5261-31-4
17464-91-4
6535-42-8
85-86-9

2475-46-9
2475-44-7
17418-58-5


 8066-05-5
63590-17-0
   58-08-2
   57-24-9
                                   8321  -  1
           Revision 0
       September 1994

-------
      Compound Name                                       CAS No.'
      Organophosphorus Compounds
      Methomyl                                            16752-77-5
      Thiofanox                                          39196-18-4
      Famphur                                               52-85-7
      Asulam                                             3337-71-1
      Dichlorvos                                            62-73-7
      Dimethoate                                            60-51-5
      Disulfoton                                          298-04-4
      Fensulfothion                                       115-90-2
      Merphos                                             150-50-5
      Methyl parathion                                    298-00-0
      Monocrotophos                                       919-44-8
      Naled                                               300-76-5
      Phorate                                             298-02-2
      Trichlorfon                                           52-68-6
      Tris-(2,3-Dibromopropyl) phosphate, (Tris-BP)        126-72-7

      Chlorinated Phenox.yacid Compounds
      Dalapon                                               75-99-0
      Dicamba                                            1918-00-9
      2,4-D                                                 94-75-7
      MCPA                                                  94-74-6
      MCPP                                               7085-19-0
      Dichlorprop                                         120-36-5
      2,4,5-T                                               93-76-5
      Silvex (2,4,5-TP)                                     93-72-1
      Dinoseb                                               88-85-7
      2,4-DB                                                94-82-6
      2,4-D, butoxyethanol ester                         1929-73-3
      2,4-D, ethylhexyl ester                            1928-43-4
      2,4,5-T,  butyl ester                                  93-79-8
      2,4,5-T,  butoxyethanol ester                       2545-59-7
      a  Chemical  Abstract Services  Registry  Number.

      1.2   This method may be applicable to  the analysis of other non-volatile
or semi volatile compounds.

      1.3   Tris-BP  has  been classified  as  a carcinogen.    Purified standard
material and stock standard solutions should be handled in  a hood.

      1.4   Method  8321   is  designed  to detect  the  chlorinated  phenoxyacid
compounds (free acid form) and their  esters  without  the  use of hydrolysis and
esterification in the extraction procedure.
                                   8321 - 2                         Revision 0
                                                                September 1994

-------
      1.5   The compounds were chosen for analysis  by HPLC/MS because they have
been designated  as  problem compounds that are hard to  analyze by traditional
chromatographic  methods  (e.g.  gas chromatography).   The  sensitivity  of this
method is dependent upon  the level of interferants within  a given matrix, and
varies  with  compound  class  and even   with  compounds  within  that  class.
Additionally,  the limit  of detection  (LOD)  is  dependent upon  the  mode  of
operation of the  mass  spectrometer.   For example,  the  LOD  for caffeine in the
selected reaction monitoring  (SRM) mode  is 45 pg  of standard  injected (10  /xL
injection), while for  Disperse Red 1  the LOD  is  180 pg.   The  LOD for caffeine
under single quadrupole scanning  is 84 pg  and is 600 pg  for  Disperse Red 1 under
similar scanning conditions.

      1.6   The  experimentally  determined limits  of  detection  (LOD)  for the
target analytes are  presented in  Tables 3, 10,  13, and  14.   For further compound
identification, MS/MS  (CAD -  collision activated dissociation)  can be used as an
optional  extension of this method.

      1.7   This  method  is restricted to  use  by or under  the  supervision  of
analysts experienced in the use of high performance liquid chromatographs/mass
spectrometers and skilled  in the interpretation of liquid chromatograms and mass
spectra.    Each  analyst must demonstrate  the  ability  to  generate  acceptable
results with this method.
2.0   SUMMARY OF METHOD

      2.1   This  method   provides   reverse   phase   high  performance  liquid
chromatographic  (RP/HPLC)  and  thermospray  (TSP)   mass  spectrometric  (MS)
conditions for the detection of the target analytes.  Quantitative analysis is
performed by TSP/MS, using an external standard approach.  Sample extracts can
be  analyzed  by direct   injection   into  the  thermospray  or  onto  a  liquid
chromatographic-thermospray interface.  A gradient  elution program is used on the
chromatograph to separate  the compounds.  Detection is achieved both by negative
ionization  (discharge  electrode)   and  positive  ionization,  with  a  single
quadrupole mass spectrometer.   Since this method is  based on an HPLC technique,
the use of ultraviolet (UV) detection is optional  on routine samples.

      2.2   Prior to  the  use  of  this method, appropriate  sample preparation
techniques must be used.

            2.2.1  Samples for  analysis of chlorinated phenoxyacid compounds are
      prepared by a modification of Method 8151 (see Sec. 7.1.2).  In general,
      one  liter of  aqueous  sample  or  fifty  grams  of solid  sample are  pH
      adjusted, extracted  with  diethyl ether,  concentrated and solvent exchanged
      to acetonitrile.

            2.2.2  Samples for  analysis of the other target analytes are prepared
      by  established  extraction techniques.   In general,  water  samples  are
      extracted at  a  neutral   pH with  methylene  chloride, using  a  separatory
      funnel  (Method  3510) or a continuous  liquid-liquid extractor  (Method
      3520).  Soxhlet  (Methods 3540/3541) or ultrasonic (Method 3550) extraction
      using methylene  chloride/acetone  (1:1)  is  used  for  solid samples.   A
                                   8321 - 3                         Revision 0
                                                                September 1994

-------
      micro-extraction technique is included  for  the extraction of Tris-BP from
      aqueous and non-aqueous matrices.

      2.3   An   optional   thermospray-mass   spectrometry/mass   spectrometry
(TS-MS/MS) confirmatory method is provided.  Confirmation is obtained by using
MS/MS collision activated dissociation (CAD)  or wire-repeller CAD.


3.0   INTERFERENCES

      3.1   Refer to Methods 3500, 3600, 8000 and 8150/8151.

      3.2   The  use  of  Florisil  Column  Cleanup  (Method  3620)  has  been
demonstrated to yield recoveries  less than 85% for some of the compounds in this
method, and is therefore not recommended for all  compounds.   Refer  to Table 2 of
Method  3620  for  recoveries  of  organophosphorus compounds  as  a  function  of
Florisil fractions.

      3.3   Compounds with  high  proton affinity  may  mask some of  the target
analytes.  Therefore, an HPLC must be used as a chromatographic separator, for
quantitative analysis.

      3.4   Analytical difficulties encountered with specific organophosphorus
compounds, as applied in this method, may include (but are not limited to) the
following:

            3.4.1  Methyl parathion shows  some minor degradation upon analysis.

            3.4.2  Naled can undergo debromination to  form dichlorvos.

            3.4.3  Merphos often  contains contamination  from merphos  oxide.
      Oxidation  of  merphos  can  occur   during   storage,  and  possibly  upon
      introduction into the mass spectrometer.

            Refer to Method 8141 for other compound problems as related to the
      various extraction methods.

      3.5   The chlorinated phenoxy acid compounds, being strong organic acids,
react  readily  with  alkaline  substances  and  may  be  lost  during  analysis.
Therefore, glassware and glass wool must be acid-rinsed, and sodium sulfate must
be acidified with sulfuric acid prior to use to avoid this possibility.

      3.6   Due to the reactivity of the chlorinated herbicides,  the standards
must  be prepared  in acetonitrile.   Methylation  will  occur  if  prepared  in
methanol.

      3.7   Solvents, reagents, glassware, and other sample processing hardware
may  yield  discrete  artifacts   or  elevated   baselines,   or  both,  causing
misinterpretation of chromatograms or spectra.   All of these materials must be
demonstrated to be free  from interferences under  the conditions of the analysis
by running reagent blanks.  Specific selection of reagents and purification of
solvents by distillation in all-glass systems may be required.
                                   8321 - 4                         Revision 0
                                                                September 1994

-------
      3.8   Interferants  co-extracted  from the sample  will  vary considerably
from source to source.  Retention times of target analytes must be verified by
using reference standards.

      3.9   The optional use of HPLC/MS/MS methods aids in the confirmation of
specific analytes.  These methods  are less  subject to chemical noise than other
mass spectrometric methods.


4.0   APPARATUS AND MATERIALS

      4.1   HPLC/MS

            4.1.1  High Performance  Liquid  Chromatograph (HPLC) - An analytical
      system  with  programmable  solvent   delivery  system  and  all  required
      accessories including 10 /xL injection  loop,  analytical  columns, purging
      gases, etc.  The solvent delivery system must be  capable, at a minimum, of
      a binary solvent  system.  The  chromatographic system  must  be capable of
      interfacing with a Mass Spectrometer (MS).

                   4.1.1.1     HPLC  Post-Column  Addition Pump - A  pump  for post-
            column  addition  should  be  used.   Ideally,  this pump  should  be a
            syringe   pump,  and  does  not  have  to  be  capable  of  solvent
            programming.

                   4.1.1.2     Recommended  HPLC  Columns  - A guard column and an
            analytical column are required.

                        4.1.1.2.1    Guard  Column  -  C18  reverse phase  guard
                   column,  10  mm x 2.6  mm  ID,  0.5  /zm frit,  or equivalent.

                        4.1.1.2.2    Analytical  Column   -  C18  reverse  phase
                   column,  100 mm  x 2 mm ID, 5 /u,m particle size of ODS-Hypersil;
                   or Ca  reversed  phase column,  100 mm x 2 mm ID,  3  urn particle
                   size of MOS2-Hypersil,  or  equivalent.

            4.1.2  HPLC/MS interface(s)

                   4.1.2.1      Micromixer - 10 /nL,  interfaces HPLC column system
            with HPLC post-column addition solvent system.

                   4.1.2.2     Interface -  Thermospray ionization interface and
            source  that  will  give  acceptable  calibration  response  for  each
            analyte of interest  at the concentration required.  The source must
            be capable of generating both  positive and negative  ions,  and have
            a discharge electrode or filament.

            4.1.3 Mass  spectrometer  system   -  A   single   quadrupole   mass
      spectrometer capable of scanning from  1  to 1000  amu.   The spectrometer
      must also be capable of scanning  from 150  to 450 amu in  1.5 sec  or  less,
      using 70  volts  (nominal) electron  energy in the positive or  negative
      electron impact modes.  In addition,  the mass spectrometer must be capable
                                   8321 -  5                         Revision 0
                                                                September 1994

-------
      of producing a calibrated mass spectrum for PEG 400,  600, or 800 (see Sec.
      5.14).

                   4.1.3.1     Optional  triple quadrupole mass  spectrometer -
            capable of generating daughter ion spectra with a collision gas in
            the second quadrupole and operation in the single quadrupole mode.

            4.1.4  Data System -  A computer system  that  allows  the continuous
      acquisition  and  storage on machine-readable  media  of all mass  spectra
      obtained throughout the duration of  the  chromatographic  program  must be
      interfaced to the mass spectrometer.  The computer must have software that
      allows any MS data file to  be searched for ions of a specified mass,  and
      such ion abundances to be plotted versus time or scan number.  This type
      of plot is defined as an Extracted Ion Current Profile (EICP).  Software
      must also be  available  that allows  integration  of  the  abundances in  any
      EICP between specified time or scan-number limits. There must be computer
      software available  to operate the specific modes of the mass spectrometer.

      4.2   HPLC  with  UV  detector  -  An  analytical  system  with  solvent
programmable  pumping  system  for  at  least  a  binary  solvent  system,  and  all
required  accessories   including  syringes,   10  /uL  injection  loop,  analytical
columns, purging gases, etc.   An  automatic injector is optional, but is useful
for multiple samples.   The columns specified in Sec. 4.1.1.2 are also used with
this system.

            4.2.1  If  the UV  detector  is to  be  used  in  tandem with  the
      thermospray  interface,   then   the   detector   cell  must  be  capable  of
      withstanding high pressures (up to  6000  psi).   However,  the  UV detector
      may be  attached  to an  HPLC independent  of  the  HPLC/TS/MS and,  in that
      case,  standard HPLC pressures are acceptable.

      4.3   Purification  Equipment for Azo Dye Standards

            4.3.1  Soxhlet extraction apparatus.

            4.3.2  Extraction  thimbles,  single  thickness,  43  x  123  mm.

            4.3.3  Filter  paper,   9.0  cm  (Whatman   qualitative   No.  1   or
      equivalent).

            4.3.4  Silica-gel  column  - 3  in.  x 8 in., packed  with Silica  gel
      (Type 60, EM reagent 70/230 mesh).

      4.4   Extraction equipment for Chlorinated Phenoxyacid Compounds

            4.4.1  Erlenmeyer flasks  - 500-mL  wide-mouth  Pyrex,  500-mL Pyrex,
      with 24/40 ground glass joint,  1000-mL pyrex.

            4.4.2  Separatory funnel  -  2000 mL.

            4.4.3  Graduated  cylinder -  1000 mL.

            4.4.4  Funnel  -  75 mm diameter.
                                   8321 - 6                         Revision 0
                                                                September 1994

-------
            4.4.5  Wrist shaker - Burrell  Model  75 or equivalent.

            4.4.6  pH meter.

      4.5   Kuderna-Danish (K-D) apparatus (optional).

            4.5.1  Concentrator tube  -  10 mi  graduated (Kontes K-570050-1025 or
      equivalent).   A  ground glass stopper  is used  to prevent  evaporation of
      extracts.

            4.5.2  Evaporation   flask -   500 mi (Kontes   K-570001-500   or
      equivalent).   Attach  to  concentrator tube  with  springs,  clamps,  or
      equivalent.

            4.5.3  Snyder column  -   Two  ball  micro  (Kontes  K-569001-0219 or
      equivalent).

            4.5.4  Springs -   1/2 in. (Kontes K-662750 or equivalent).

      4.6   Disposable serological pipets -  5 ml x 1/10, 5.5 mm ID.

      4.7   Collection  tube  - 15 mi  conical,  graduated (Kimble No.  45165 or
equivalent).

      4.8   Vials -  5 ml conical, glass, with Teflon lined screw-caps or crimp
tops.

      4.9   Glass wool - Supelco No. 2-0411  or equivalent.

      4.10  Microsyringes - 100 jil, 50 jzL, 10 p.i (Hamilton 701  N or equivalent),
and 50 ni  (Blunted,  Hamilton  705SNR  or equivalent).

      4.11  Rotary evaporator  - Equipped with 1000 ml  receiving flask.

      4.12  Balances - Analytical, 0.0001 g,  Top-loading, 0.01 g.

      4.13  Volumetric flasks, Class A - 10 ml to 1000 ml.

      4.14  Graduated cylinder - 100 ml.

      4.15  Separatory funnel  - 250 ml.


5.0   REAGENTS

      5.1   Reagent  grade inorganic  chemicals  shall  be used  in  all  tests.
Unless otherwise indicated, it  is intended  that all reagents shall conform to the
specifications of the Committee on Analytical Reagents  of the American Chemical
Society, where  such  specifications  are available.  Other grades may  be  used,
provided it is first  ascertained that the reagent  is of sufficiently high purity
to permit its use without lessening the accuracy of the determination.
                                   8321 - 7                         Revision 0
                                                                September 1994

-------
      5.2   Organic free reagent water.  All  references to water in  this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Sodium sulfate  (granular,  anhydrous),  Na2S04. Purify by heating  at
400°C for 4 hours in  a shallow tray, or by precleaning the sodium sulfate with
methylene chloride..

      5.4   Ammonium acetate, NH4OOCCH3, solution (0.1 M).  Filter through a 0.45
micron membrane filter (Millipore HA or equivalent).

      5.5   Acetic acid, CH3C02H

      5.6   Sulfuric acid solution

            5.6.1  ((1:1)  (v/v)) - Slowly add 50 ml H2S04  (sp. gr. 1.84) to   50
      mL of water.

            5.6.2  ((1:3)  (v/v)) - slowly add 25 ml H2S04  (sp. gr. 1.84) to   75
      ml of water.

      5.7   Argon gas, 99+%  pure.

      5.8   Solvents

            5.8.1  Methylene chloride,  CH2C12 - Pesticide quality or  equivalent.

            5.8.2  Toluene,  C6H5CH3 - Pesticide quality or equivalent.

            5.8.3  Acetone,  CH3COCH3 -  Pesticide quality  or equivalent.

            5.8.4  Diethyl  Ether, C2H5OC2H5 -  Pesticide quality or  equivalent.
      Must  be  free of  peroxides as  indicated by  test  strips  (EM Quant,   or
      equivalent).  Procedures  for  removal of peroxides are provided with the
      test strips.  After cleanup, 20 ml of ethyl alcohol preservative must  be
      added to each liter of ether.

            5.8.5  Methanol, CH3OH - HPLC quality  or equivalent.

            5.8.6  Acetonitrile, CH3CN  -  HPLC  quality or equivalent.

            5.8.7  Ethyl  acetate CH3C02C2H5 -  Pesticide quality or equivalent.

      5.9   Standard Materials - pure  standard materials  or certified solutions
of each  analyte targeted for analysis.   Disperse azo dyes must be  purified before
use according to Sec. 5.10.

      5.10  Disperse Azo Dye Purification

            5.10.1      Two  procedures are involved.   The  first  step  is the
      Soxhlet extraction of the  dye for 24 hours with toluene  and evaporation  of
      the liquid extract to  dryness,  using a  rotary evaporator.  The solid  is
      then recrystallized from  toluene,  and  dried in an oven at approximately
      100°C.   If  this  step  does  not   give  the  required  purity,  column

                                   8321  - 8                         Revision 0
                                                                 September 1994

-------
      chromatography  should  be employed.   Load  the solid  onto  a 3  x  8 inch
      silica gel column  (Sec.  4.3.4),  and  elute  with  diethyl  ether.   Separate
      impurities chromatographically, and collect the major dye fraction.

      5.11  Stock  standard  solutions  - Can  be  prepared  from  pure standard
materials or can be purchased as certified solutions.

            5.11.1       Prepare stock standard solutions  by accurately weighing
      0.0100 g  of  pure material.   Dissolve the  material  in  methanol or other
      suitable  solvent  (e.g.  prepare Tris-BP  in  ethyl  acetate),  and  dilute to
      known volume in a volumetric flask.

            NOTE:  Due to  the reactivity  of  the chlorinated  herbicides,  the
                   standards  must be prepared in acetonitrile.  Methylation will
                   occur if prepared  in methanol.

            If  compound  purity  is  certified  at 96%  or greater,  the weight can
      be used without correction  to  calculate the  concentration  of  the stock
      standard.    Commercially  prepared stock  standards can  be  used  at  any
      concentration  if  they  are  certified  by  the   manufacturer  or  by  an
      independent source.

            5.11.2      Transfer the stock standard solutions into glass vials
      with Teflon lined screw-caps  or crimp-tops.  Store at 4°C and protect from
      light.  Stock standard  solutions should be checked frequently for signs of
      degradation or  evaporation, especially just prior to preparing calibration
      standards.

      5.12  Calibration  standards  - A  minimum of five concentrations for each
parameter of interest should  be prepared through dilution  of the stock standards
with methanol  (or other suitable solvent).   One of these  concentrations should
be near, but above, the MDL.   The remaining concentrations should correspond to
the expected range  of concentrations found in real samples, or should define the
working range of the  HPLC-UV/VIS or HPLC-TSP/MS.   Calibration standards must be
replaced after one or two months,  or sooner if comparison with check standards
indicates a problem.

      5.13  Surrogate standards -  The analyst  should monitor the performance of
the extraction,  cleanup (when  used),  and analytical  system,  along  with  the
effectiveness of the  method in dealing  with each  sample matrix, by spiking each
sample, standard, and blank with one  or two surrogates (e.g.,  organophosphorus
or chlorinated phenoxyacid  compounds  not expected to be present in the sample).

      5.14  HPLC/MS tuning  standard - Polyethylene glycol  400 (PEG-400), PEG-600
or PEG-800.  Dilute  to 10  percent (v/v) in methanol.  Dependent  upon analyte
molecular weight range: m.w. < 500  amu,  use PEG-400; m.w. > 500 amu, use PEG-600,
or PEG-800.
6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   See the  introductory  material  to this Chapter,  Organic Analytes,
Sec. 4.1.
                                   8321 - 9                         Revision 0
                                                                September 1994

-------
7.0   PROCEDURE

      7.1   Sample preparation - Samples for analysis of disperse azo dyes and
organophosphorus compounds must be prepared by one of the following methods prior
to HPLC/MS analysis:

      Matrix                                    Methods

      Water                                     3510,  3520
      Soil/sediment                             3540,  3541, 3550
      Waste                                     3540,  3541, 3550, 3580

      Samples for the analysis of Tris-(2,3-dibromopropyl )phosphate in wastewater
must be prepared according to Sec. 7.1.1 prior to HPLC/MS analysis.  Samples for
the analysis  of chlorinated phenoxyacid compounds and  their  esters  should be
prepared according to Sec. 7.1.2 prior to HPLC/MS analysis.

            7.1.1  Microextraction  for Tris-BP:

                   7.1.1.1     Solid Samples

                        7.1.1.1.1   Weigh  a 1 gram  portion  of the sample into
                   a  tared beaker.    If  the  sample  appears  moist,  add  an
                   equivalent amount of anhydrous  sodium sulfate and mix well.
                   Add 100 /iL of Tris-BP (approximate concentration 1000 mg/L)
                   to the  sample selected for spiking; the amount added should
                   result   in a  final  concentration  of  100  ng//iL in the  1 mL
                   extract.

                        7.1.1.1.2   Remove the glass wool plug from a disposable
                   serological  pipet.    Insert  a  1  cm  plug  of  clean  silane
                   treated glass wool  to the bottom  (narrow end) of the pipet.
                   Pack 2  cm of  anhydrous  sodium  sulfate onto  the  top  of the
                   glass  wool.   Wash  pipet  and contents   with  3  - 5  mL of
                   methanol.

                        7.1.1.1.3   Pack  the sample  into the  pipet prepared
                   according to  Sec. 7.1.1.1.2.  If packing material has dried,
                   wet with a few mL  of  methanol  first,  then  pack sample into
                   the  pipet.

                        7.1.1.1.4   Extract  the sample with  3  mL of methanol
                   followed by 4 mL of 50%  (v/v) methanol/methylene chloride
                   (rinse   the  sample  beaker  with each  volume  of  extraction
                   solvent  prior  to  adding it to  the  pipet  containing  the
                   sample).  Collect  the extract in  a 15 mL graduated  glass
                   tube.

                        7.1.1.1.5   Evaporate  the extract  to 1 mL  using  the
                   nitrogen blowdown  technique  (Sec.  7.1.1.1.6).   Record  the
                   volume.   It  may not  be  possible  to  evaporate  some  sludge
                   samples to a  reasonable concentration.
                                   8321  -  10                        Revision 0
                                                                September 1994

-------
      7.1.1.1.6   Nitrogen  Slowdown  Technique

            7.1.1.1.6.1       Place  the  concentrator tube  in
      a warm water bath (approximately 35°C) and evaporate the
      solvent  volume to  the  required level  using a gentle
      stream' of  clean,  dry  nitrogen  (filtered  through   a
      column of activated carbon).

            CAUTION:     Do   not  use   plasticized  tubing
                         between  the  carbon  trap  and  the
                         sample.

            7.1.1.1.6.2       The internal wall of  the  tube
      must  be  rinsed  down  several  times  with  methylene
      chloride  during the operation.   During evaporation, the
      solvent level  in the tube must  be positioned  to prevent
      water from  condensing  into  the   sample  (i.e.,  the
      solvent  level  should  be below the  level  of the water
      bath).       Under  normal   operating  conditions,  the
      extract  should not  be allowed  to become dry.   Proceed
      to  Sec.  7.1.1.1.7.

      7.1.1.1.7   Transfer  the extract  to a glass vial  with
a Teflon lined screw-cap  or crimp-top and  store refrigerated
at 4°C.   Proceed with HPLC  analysis.

      7.1.1.1.8   Determination  of  percent  dry  weight  -  In
certain  cases,  sample results  are  desired based  on a dry
weight basis.   When such data  are  desired, or required,   a
portion  of sample  for  this determination should  be  weighed
out  at  the same  time as  the  portion  used for  analytical
determination.

      WARNING:     The drying  oven should  be contained  in  a
                  hood  or  vented.    Significant  laboratory
                  contamination   may result  from  drying   a
                  heavily  contaminated    hazardous  waste
                  sample.

      7.1.1.1.9   Immediately  after  weighing the  sample for
extraction, weigh 5-10 g  of the sample into a tared crucible.
Determine the % dry weight  of  the sample by  drying overnight
at 105°C.  Allow to  cool  in a  desiccator before weighing:

      % dry weight = q  of dry  sample x 100
                       g  of sample
                8321  -  11                         Revision 0
                                             September 1994

-------
             7.1.1.2     Aqueous Samples

                  7.1.1.2.1   Using a 100 ml graduated cylinder, measure
             100 ml  of  sample  and  transfer it  to  a 250  ml separatory
             funnel.  Add  200  juL of Tris-BP  (approximate concentration
             1000 mg/L)  to  the  sample  selected  for  spiking;  the amount
             added should  result in a final concentration of 200 ng/juL in
             the 1 ml extract.

                  7.1.1.2.2   Add  10  ml  of methylene  chloride  to  the
             separatory  funnel.   Seal   and  shake the  separatory funnel
             three  times,   approximately  30  seconds  each  time,  with
             periodic venting to release excess  pressure.  NOTE: Methylene
             chloride creates   excessive  pressure  rapidly;  therefore,
             initial  venting   should   be  done  immediately  after  the
             separatory  funnel has been sealed and shaken once.  Methylene
             chloride is  a suspected  carcinogen,  use  necessary safety
             precautions.

                  7.1.1.2.3   Allow the organic layer to separate from the
             water phase for a  minimum  of 10 minutes.   If the emulsion
             interface between  layers  is more than one-third the size of
             the  solvent   layer,  the   analyst   must  employ  mechanical
             techniques  to  complete phase  separation.   See  Sec.  7.5,
             Method 3510.

                  7.1.1.2.4   Collect  the extract in  a  15 ml graduated
             glass tube. Proceed as  in  Sec.  7.1.1.1.5.

      7.1.2  Extraction  for chlorinated phenoxyacid compounds - Preparation
of soil, sediment, and other solid samples must  follow Method 8151, with
the exception of no hydrolysis or esterification.  Sec. 7.1.2.1 presents
an outline  of  the procedure with the appropriate  changes necessary for
determination  by Method 8321.   Sec.  7.1.2.2  describes  the extraction
procedure for aqueous samples.

             7.1.2.1     Extraction  of  solid samples

                  7.1.2.1.1   Add 50 g  of soil/sediment sample to a 500
             ml,  wide  mouth   Erlenmeyer.    Add spiking  solutions  if
             required, mix well  and allow to stand for 15 minutes.  Add 50
             ml  of  organic-free  reagent water  and stir  for  30 minutes.
             Determine the pH  of the sample  with  a glass electrode and pH
             meter,  while stirring.   Adjust  the  pH to 2 with  cold  H2S04
             (1:1) and monitor the pH  for 15 minutes, with stirring.  If
             necessary,  add additional  H2S04 until the  pH  remains  at  2.

                  7.1.2.1.2   Add 20 mL of  acetone to the  flask, and mix
             the contents  with the wrist shaker for 20 minutes.   Add 80 ml
             of diethyl  ether to the same flask, and shake again for 20
             minutes.   Decant  the  extract  and  measure  the  volume of
             solvent recovered.
                             8321 - 12                        Revision 0
                                                          September  1994

-------
      7.1.2.1.3    Extract the sample twice more using 20 ml
of  acetone  followed  by  80  ml  of  diethyl  ether.   After
addition of each solvent, the mixture should be shaken with
the wrist shaker for 10 minutes and the acetone-ether extract
decanted.

      7.1.2.1.4    After  the  third extraction, the volume of
extract  recovered  should be  at least 75%  of the volume of
added  solvent.     If  this   is   not  the   case,   additional
extractions may  be necessary.  Combine the extracts in  a 2000
ml separatory funnel  containing 250 ml of  reagent  water.  If
an emulsion forms,  slowly add 5 g of acidified sodium sulfate
(anhydrous)  until  the solvent-water  mixture  separates.   A
quantity of acidified sodium sulfate  equal to the weight of
the sample may be added, if necessary.

      7.1.2.1.5    Check  the pH  of the extract.   If it  is not
at or below pH 2, add more concentrated HC1  until the extract
is stabilized at the  desired  pH.   Gently mix  the contents of
the separatory funnel for 1  minute and  allow the layers to
separate.  Collect the aqueous phase  in a  clean beaker, and
the  extract phase  (top  layer)  in  a 500  ml  ground-glass
Erlenmeyer  flask.   Place the  aqueous phase back  into the
separatory  funnel  and  re-extract  using  25  ml  of diethyl
ether. Allow the layers to separate  and discard the aqueous
layer. Combine the ether extracts in the  500 ml  Erlenmeyer
flask.

      7.1.2.1.6    Add  45 - 50  g  acidified anhydrous  sodium
sulfate to the combined ether extracts.  Allow the  extract to
remain in contact with the sodium  sulfate  for approximately
2 hours.

      NOTE:  The drying step is very  critical.  Any  moisture
            remaining  in  the   ether will   result  in  low
            recoveries.   The amount  of sodium sulfate used is
            adequate  if  some   free  flowing  crystals  are
            visible when swirling the flask.  If all  of the
            sodium sulfate solidifies in  a cake,  add a few
            additional grams  of acidified  sodium sulfate and
            again test by swirling.  The 2 hour drying time is
            a minimum;  however,   the extracts  may  be held
            overnight in contact with the sodium sulfate.

      7.1.2.1.7    Transfer the ether extract, through a  funnel
plugged with acid-washed glass  wool, into  a 500 ml K-D flask
equipped with a  10 ml  concentrator tube.  Use a glass  rod to
crush caked sodium sulfate during the transfer.   Rinse the
Erlenmeyer flask and column with 20-30 ml of diethyl  ether to
complete the quantitative transfer.  Reduce the volume  of the
extract using the macro  K-D  technique (Sec. 7.1.2.1.8).

      7.1.2.1.8   Add  one or  two  clean boiling  chips to the
flask and attach a three ball  macro-Snyder column.   Prewet

                8321  - 13                        Revision 0
                                             September 1994

-------
the Snyder column by  adding  about  1  ml of diethyl ether to
the top.  Place the apparatus on a  hot water bath (60°-65°C)
so that the concentrator  tube  is  partially immersed in the
hot water and the entire lower rounded surface of the flask
is bathed  in  vapor.    Adjust the vertical  position  of the
apparatus and  the water temperature, as required,  to complete
the concentration in  15-20 minutes.   At  the proper rate of
distillation the balls of the column will actively chatter,
but the chambers will  not flood.  When the apparent volume of
liquid reaches 5 ml,  remove the  K-D apparatus from the water
bath and allow it  to  drain and cool for at least  10 minutes.

      7.1.2.1.9    Exchange  the  solvent  of  the  extract to
acetonitrile by quantitatively transferring  the extract with
acetonitrile to a blow-down  apparatus.  Add a total of 5 ml
acetonitrile.   Reduce  the extract  volume according to Sec.
7.1.1.1.6, and adjust the final  volume to 1 ml.

7.1.2.2     Preparation of aqueous samples

      7.1.2.2.1    Using a 1000 mL graduated cylinder, measure
1 liter (nominal)  of  sample,  record the sample volume to the
nearest 5 mL,  and  transfer  it to a  separatory  funnel.   If
high concentrations are anticipated,  a smaller volume may be
used and  then diluted with organic-free  reagent water to  1
liter. Adjust  the  pH  to less than 2  with sulfuric  acid (1:1).

      7.1.2.2.2    Add 150  mL  of diethyl ether to the sample
bottle, seal,  and shake for  30  seconds to rinse the walls.
Transfer  the  solvent  wash   to  the  separatory  funnel  and
extract the sample by shaking the funnel  for 2 minutes with
periodic  venting  to  release excess  pressure.    Allow the
organic layer to separate  from  the water  layer for a minimum
of 10  minutes.  If  the emulsion  interface between layers is
more  than one-third  the  size  of the solvent  layer,  the
analyst must  employ  mechanical  techniques  to  complete the
phase  separation.  The optimum  technique depends  upon the
sample, and may include stirring, filtration of the emulsion
through  glass  wool,  centrifugation,  or  other  physical
methods.  Drain the aqueous phase into a  1000 ml  Erlenmeyer
flask.

      7.1.2.2.3    Repeat the  extraction two  more  times using
100 mL of diethyl ether each time.  Combine the  extracts in
a  500  mL  Erlenmeyer  flask.   (Rinse  the  1000  mL flask  with
each  additional  aliquot  of  extracting  solvent  to  make  a
quantitative transfer.)

      7.1.2.2.4    Proceed  to  Sec.  7.1.2.1.6   (drying,  K-D
concentration,   solvent   exchange,   and   final   volume
adjustment).
                8321 - 14                         Revision  0
                                             September  1994

-------
      7.2   Prior to HPLC analysis, the extraction  solvent must be exchanged to
methanol or acetonitrile (Sec.  7.1.2.1.9).   The exchange  is performed using the
K-D procedures listed in all of the extraction methods.

      7.3   HPLC Chromatographic Conditions:

            7.3.1  Analyte-specific Chromatographic  conditions  are shown  in
      Table 1.  Chromatographic conditions which are not analyte-specific  are as
      follows:

            Flow rate:                     0.4 mL/min
            Post-column mobile phase:      0.1 M  ammonium acetate (1% methanol)
                                           (0.1   M   ammonium   acetate   for
                                           phenoxyacid compounds)
            Post-column flow rate:         0.8 mL/min

            7.3.2  If there  is  a Chromatographic problem from compound retention
      when  analyzing for disperse azo  dyes,  organophosphorus  compounds,  or
      Tris-(2,3-dibromopropyl)phosphate,  a  2%  constant  flow  of  methylene
      chloride may be applied  as  needed.   Methylene chloride/aqueous methanol
      solutions must be used with  caution  as  HPLC  eluants.   Acetic acid (1%),
      another  mobile phase  modifier,  can  be  used  with compounds  with acid
      functional groups.

            7.3.3  A  total  flow  rate  of 1.0  to 1.5  mL/min  is necessary  to
      maintain thermospray ionization.

            7.3.4  Retention  times  for  organophosphorus   compounds  on  the
      specified analytical column are presented in Table 9.

      7.4   Recommended HPLC/Thermospray/MS operating conditions:

            7.4.1  Positive  Ionization mode

            Repeller (wire or  plate, optional):  170   to  250   v   (sensitivity
      optimized). See Figure 2 for schematic of source with  wire repeller.

            Mass range: 150 to 450 amu (compound dependent,  expect 1 to  18 amu
                        higher than molecular weight of  the compound).
            Scan time:  1.50 sec/scan.

            7.4.2  Negative  Ionization mode

            Discharge electrode:     on
            Filament:               off
            Mass Range:             135 to 450 amu
            Scan time:              1.50 sec/scan.

            7.4.3  Thermospray  temperatures:
            Vaporizer control  110°C  to  130°C.
            Vaporizer tip      200°C  to  215°C.
                                   8321  -  15                         Revision 0
                                                                September 1994

-------
      Jet                210°C  to  220°C.
      Source block       230°C  to 265°C.  (Some compounds  may degrade in
                         the source  block  at  higher  temperatures,  the
                         operator  should   use   knowledge  of  chemical
                         properties    to    estimate    proper    source
                         temperature).

      7.4.4  Sample injection  volume:  20  /xL is  necessary  in  order to
overfill the 10 pi injection loop.  If  solids are  present  in  the extract,
allow them to settle or centrifuge the  extract and withdraw the injection
volume from the clear layer.

7.5   Calibration:

      7.5.1  Thermospray/MS system -  Must be  hardware-tuned on quadrupole
1 (and quadrupole 3 for triple  quadrupoles)  for accurate mass assignment,
sensitivity, and  resolution.    This  is  accomplished  using  polyethylene
glycol (PEG) 400,  600,  or 800 (see Sec.  5.14) which have average molecular
weights of 400, 600, and 800, respectively.  A mixture of these PEGs can
be made such that it will  approximate the expected working mass range for
the  analyses.   Use PEG  400  for  analysis  of  chlorinated  phenoxyacid
compounds.    The  PEG  is   introduced  via   the  thermospray  interface,
circumventing the HPLC.

             7.5.1.1     The mass calibration parameters are as follows:

             for PEG 400 and 600                 for PEG 800
             Mass range:  15 to  765 amu            Mass range:  15 to 900 amu
             Scan time: 5.00 sec/scan            Scan time: 5.00 sec/scan

             Approximately  100  scans  should  be   acquired,  with  2  to  3
      injections made.  The scan  with  the  best  fit  to the accurate mass
      table (see Tables 7 and 8) should be  used as the calibration table.

             7.5.1.2     The low mass range from  15  to  100 amu is  covered
      by the ions from the  ammonium acetate buffer used in  the thermospray
      process:  NH4+  (18 amu),  NH4+H20 (36), CH3OHNH4 + (50) (methanol),
      or  CH3CNNH4+  (59)  (acetonitrile),  and  CH3COOHNH4+  (78)   (acetic
      acid). The appearance of  the m/z 50 or 59 ion depends  upon  the use
      of methanol  or  acetonitrile as the  organic modifier.   The higher
      mass  range  is covered by the  ammonium ion  adducts  of the  various
      ethylene   glycols    (e.g.   H(OCH2CH2)nOH   where  n=4,   gives  the
      H(OCH2CH2)4OH-NH4+ ion at  m/z 212).

      7.5.2  Liquid Chromatograph

             7.5.2.1     Prepare calibration standards as outlined in Sec.
      5.12.

             7.5.2.2     Choose  the proper ionization   conditions,  as
      outlined  in  Sec.  7.4.   Inject each  calibration standard onto the
      HPLC,  using the chromatographic conditions outlined  in  Table 1.
      Calculate the area under  the curve  for the mass chromatogram of each


                             8321  -  16                         Revision 0
                                                          September  1994

-------
quantitation ion.  For example, Table  9  lists  the retention times
and  the  major   ions  (>5%)  present  in  the  positive  ionization
thermospray  single  quadrupole  spectra  of  the  organophosphorus
compounds.  In most cases the (M+H)+  and (M+NH4)+ adduct ions are
the only  ions of significant abundance.   Plot  these  ions  as area
response versus the amount  injected.   The points  should  fall on a
straight line, with a correlation coefficient of at least 0.99 (0.97
for chlorinated phenoxyacid analytes).

       7.5.2.3      If  HPLC-UV  detection  is   also   being  used,
calibrate  the  instrument  by preparing  calibration   standards  as
outlined in Sec.  5.12, and  injecting each calibration standard onto
the HPLC using the chromatographic conditions outlined in Table 1.
Integrate  the  area under the full  chromatographic peak  for each
concentration.   Quantitation by HPLC-UV  may  be  preferred if it is
known that sample  interference and/or  analyte  coelution  are not a
problem.

       7.5.2.4      For  the  methods  specified in  Sec.  7.5.2.2  and
7.5.2.3,  the  retention  time of  the   chromatographic  peak  is  an
important variable in analyte identification.  Therefore, the ratio
of the retention  time of  the sample analyte to the standard analyte
should be 1.0 - 0.1.

       7.5.2.5      The concentration of the  sample analyte will be
determined by  using the calibration   curves  determined  in  Sees.
7.5.2.2 and 7.5.2.3.  These  calibration curves must be generated on
the  same  day  as  each  sample is  analyzed.    At  least  duplicate
determinations should be made for  each  sample extract.   Samples
whose concentrations exceed  the standard calibration  range should
be diluted to fall within the range.

       7.5.2.6      Refer  to Method 8000 for  further information on
calculations.

       7.5.2.7      Precision can also  be  calculated  from the ratio
of response (area) to the  amount  injected; this is defined as the
calibration factor (CF)  for each  standard concentration.   If the
percent relative standard deviation (%RSD) of  the CF  is  less than
20 percent over the working  range, linearity through the origin can
be assumed, and the average  calibration factor can be used in place
of a  calibration  curve.   The CF  and %RSD  can be calculated  as
follows:

       CF  = Total  Area of Peak/Mass  injected  (ng)

       %RSD = SD/CF x 100

where:

       SD  = Standard  deviation between  CFs

       CF  = Average CF


                      8321  -  17                         Revision 0
                                                    September 1994

-------
      7.6   Sample Analysis

            7.6.1  Once the  LC/MS system has been calibrated  as outlined in Sec.
      7.5, it is ready for sample  analysis.   It  is recommended that the samples
      initially be analyzed in the negative ionization mode.  If low levels of
      compounds are suspected, then the samples should also be screened in the
      positive ionization mode.

                   7.6.1.1     A  blank 20  p,L  injection  (methanol)  must  be
            analyzed after the standard(s) analyses, in order to determine any
            residual contamination of the Thermospray/HPLC/MS system.

                   7.6.1.2     Take a 20 /xL aliquot of the sample extract from
            Sec. 7.4.4.   Start  the  HPLC  gradient  elution,  load  and inject the
            sample  aliquot,  and   start  the  mass  spectrometer  data  system
            analysis.

      7.7   Calculations

            7.7.1  Using  the  external  standard calibration  procedure (Method
      8000), determine the identity and quantity of each component peak in the
      sample reconstructed ion chromatogram which corresponds to the compounds
      used  for  calibration  processes.    See  Method  8000  for  calculation
      equations.

            7.7.2  The retention time of the chromatographic peak  is an important
      parameter  for the  identity  of  the analyte.   However,  because  matrix
      interferences can change chromatographic column conditions, the retention
      times are  not as  significant,  and the  mass spectra  confirmations are
      important criteria  for analyte identification.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter  One  and  Method  8000 for  specific quality control
procedures.

      8.2   Tables 4, 5, 6,  11,  12, and 15 indicate the  single operator accuracy
and precision for this  method.  Compare the results obtained  with the results in
the tables to determine if the data quality is acceptable.  Tables 4, 5,  and 6
provide  single  lab  data  for Disperse  Red  1,  Table  11  with  organophoshorus
pesticides, Table 12 with Tris-BP and Table 15 with  chlorophenoxyacid herbicides.

            8.2.1  If recovery is  not acceptable,  check the  following:

                   8.2.1.1     Check to be sure  that there are no errors in the
            calculations, surrogate solutions or internal  standards.  If errors
            are found,  recalculate the data accordingly.

                   8.2.1.2     Check instrument performance.  If an instrument
            performance   problem   is   identified,   correct   the  problem  and
            re-analyze  the extract.
                                   8321  -  18                        Revision 0
                                                                September 1994

-------
                   8.2.1.3     If no problem is found,  re-extract and re-analyze
            the sample.

                   8.2.1.4     If, upon re-analysis,  the recovery is again not
            within limits, flag the data as "estimated concentration".

      8.3   Instrument  performance  -  Check  the  performance of  the  entire
analytical  system daily using  data gathered from analyses of blanks, standards,
and replicate samples.

            8.3.1  See Sec. 7.5.2.7 for required HPLC/MS  parameters for standard
      calibration curve %RSD limits.

            8.3.2  See Sec.  7.5.2.4 regarding  retention  time  window QC limits.

            8.3.3  If any of  the chromatographic QC  limits  are not  met,  the
      analyst should examine the LC system for:

            •      Leaks,
            •      Proper  pressure delivery,
            •      A dirty guard column;  may need  replacing or repacking,  and
            •      Possible  partial thermospray plugging.

            Any of the above items will necessitate shutting  down the HPLC/TSP
      system,   making repairs  and/or  replacements,  and  then restarting  the
      analyses.  The  calibration standard should be reanalyzed before any sample
      analyses, as described in Sec.  7.5.

            8.3.4   The   experience    of  the   analyst   performing   liquid
      chromatography is invaluable to the success of the method.  Each day that
      analysis is performed, the daily  calibration standard should be evaluated
      to determine if the  chromatographic system is operating properly.  If any
      changes are made to the  system (e.g. column  change), the system must be
      recalibrated.

      8.4   Optional Thermospray HPLC/MS/MS confirmation

            8.4.1  With respect to this method, MS/MS shall  be defined as the
      daughter ion collision activated  dissociation acquisition with quadrupole
      one set on one mass (parent  ion),  quadrupole two  pressurized with argon
      and with a higher offset  voltage  than normal, and quadrupole three set to
      scan  desired mass range.

            8.4.2  Since the thermospray process often generates only one or two
      ions  per compound,  the use of MS/MS is a more specific mode of operation,
      yielding molecular structural information.   In  this mode, fast screening
      of samples can  be accomplished through direct injection  of the sample into
      the thermospray.

            8.4.3  For MS/MS experiments,  the first quadrupole should be set to
      the protonated molecule  or ammoniated adduct of the analyte of interest.
      The third quadrupole should be  set to scan  from 30 amu  to just above the
      mass  region of the protonated molecule.


                                   8321  -  19                         Revision 0
                                                                September 1994

-------
            8.4.4  The  collision  gas pressure (Ar) should be set  at  about  1.0
      mTorr and the collision energy at 20 eV.   If  these parameters fail to give
      considerable fragmentation, they may  be raised above these  settings  to
      create more and stronger collisions.

            8.4.5  For  analytical determinations, the base peak of the collision
      spectrum shall  be taken as the quantification ion.  For extra specificity,
      a second ion should be chosen as a  backup quantification  ion.

            8.4.6  Generate  a  calibration curve  as outlined  in Sec.  7.5.2.

            8.4.7  For  analytical determinations, calibration blanks must be run
      in the  MS/MS  mode to  determine  specific   ion  interferences.   If  no
      calibration  blanks  are  available,  chromatographic  separation must  be
      performed  to assure  no interferences  at  specific  masses.    For  fast
      screening, the MS/MS  spectra of the  standard  and  the analyte  could  be
      compared and the  ratios of the three major (most intense)  ions examined.
      These  ratios  should  be  approximately  the  same,   unless   there  is  an
      interference.    If an   interference  appears,  chromatography  must  be
      utilized.

            8.4.8  For  unknown concentrations,  the total area of the quantitation
      ion(s) is calculated and the calibration curves generated  as in Sec.  7.5
      are used to attain an  injected weight number.

            8.4.9  MS/MS  techniques  can  also  be   used  to   perform  structural
      analysis on ions  represented by  unassigned m/z ratios.  The procedure for
      compounds of unknown structures is  to set up a CAD  experiment on the ion
      of interest.  The spectrum generated from this experiment will reflect the
      structure of the compound by  its fragmentation  pattern.  A trained  mass
      spectroscopist  and  some history  of  the  sample are  usually  needed  to
      interpret  the  spectrum.  (CAD experiments  on  actual  standards of  the
      expected  compound  are  necessary  for  confirmation  or denial  of  that
      substance.)

      8.5   Optional  wire-repeller CAD confirmation

            8.5.1  See  Figure 3 for the correct position of the wire-repeller in
      the thermospray source block.

            8.5.2  Once the wire-repeller is inserted  into the thermospray flow,
      the voltage can be  increased to approximately 500 - 700 v.  Enough voltage
      is necessary to  create  fragment  ions,  but not so  much   that  shorting
      occurs.

            8.5.3  Continue  as outlined in  Sec.  7.6.
9.0   METHOD PERFORMANCE

      9.1   Single operator accuracy and precision studies have been conducted
using spiked sediment, wastewater, sludge,  and water samples.   The results are
presented in Tables 4, 5,  6,  11,  12,  and  15.  Tables 4, 5, and 6 provide single


                                   8321  - 20                         Revision 0
                                                                September 1994

-------
lab data for Disperse Red 1, Table 11 for organophoshorus pesticides, Table 12
for Tris-BP and Table 15 with chlorophenoxyacid herbicides.

      9.2   LODs should be calculated for the known analytes, on each instrument
to be used.  Tables 3, 10, and 13 list limits of detection  (LOD) and/or estimated
quantitation limits (EQL) that are typical  with this method.

            9.2.1  The LODs presented in this method were calculated by analyzing
      three  replicates  of   four   standard  concentrations,  with  the  lowest
      concentration  being near  the  instrument  detection  limit.    A  linear
      regression was  performed on  the data  set  to  calculate  the  slope  and
      intercept. Three times the standard deviation (3a)  of the lowest standard
      amount, along with the calculated slope and intercept,  were used to find
      the LOD.  The LOD was   not calculated using the specifications in Chapter
      One,  but according to   the ACS guidelines specified  in Reference 4.

            9.2.2  Table 17  presents a comparison  of the  LODs  from Method 8151
      and Method 8321 for the chlorinated phenoxyacid compounds.

      9.3   Table  16  presents multilaboratory  accuracy and precision data for
the chlorinated phenoxyacid  herbicides.  The data  summary is based on data from
three  laboratories   that   analyzed  duplicate  solvent   solutions   at  each
concentration specified in the Table.


10.0  REFERENCES

1.    Voyksner, R.D.; Haney,  C.A.   "Optimization and Application of Thermospray
      High-Performance  Liquid  Chromatography/Mass Spectrometry"; Anal.  Chem.
      1985, 57, 991-996.

2.    Blakley,  C.R.;   Vestal,  M.L.     "Thermospray  Interface   for  Liquid
      Chromatography/Mass Spectrometry"; Anal. Chem. 1983,  55,  750-754.

3.    Taylor, V.; Mickey,  D.  M., Marsden, P.  J.  "Single  Laboratory Validation of
      EPA Method 8140";  EPA-600/4-87/009, U.S.  Environmental Protection Agency,
      Las Vegas, NV, 1987, 144 pp.

4.    "Guidelines  for  Data Acquisition  and Data   Quality   Evaluation  in
      Environmental Chemistry"; Anal. Chem.  1980,  5_2,  2242-2249.

5.    Betowski, L.  D.; Jones, T. L.  "The Analysis of Organophosphorus Pesticide
      Samples by HPLC/MS and  HPLC/MS/MS"; Environmental Science and Technology.
      1988,

8.    EPA:   2nd Annual Report on Carcinogens,  NTP 81-43, Dec. 1981, pp. 236-237.

9.    Blum, A.; Ames, B. N.   Science 195, 1977, 17.

10.   Zweidinger,   R.  A.; Cooper,  S.  D.;  Pellazari,   E.  D.,  Measurements  of
      Organic Pollutants in  Water  and Wastewater,  ASTM 686.

11.   Cremlyn, R.   Pesticides:  Preparation and mode of Action; John  Wiley and
      Sons:  Chichester, 1978; p.  142.

                                   8321  - 21                         Revision 0
                                                                September 1994

-------
12.    Cotterill,   E.  6.;  Byast,  T.  H.   "HPLC   of  Pesticide   Residues  in
      Environmental   Samples."     in  Liquid  Chromatography  in  Environmental
      Analysis; Laurence, J.  F.,  Ed.; Humana Press:  Clifton,  NJ,  1984.

13.    Voyksner, R. I).  "Thermospray HPLC/MS for Monitoring the Environment."  In
      Applications of New Mass Spectrometry  Techniques in Pesticide Chemistry;
      Rosen, J. D.,  Ed.,  John Wiley and Sons:   New York,  1987.

14.    Yinon, J.;  Jones,  T. L.;  Betowski,  L.  D.  Rap. Comm. Mass  Spectrom. 1989,
      3,  38.

15.    Shore, F. L.;  Amick, E. N.,  Pan, S. T.,  Gurka,  D.  F.  "Single Laboratory
      Validation of EPA Method 8150 for the  Analysis of Chlorinated Herbicides
      in   Hazardous  Waste";  EPA/600/4-85/060,   U.S.   Environmental  Protection
      Agency, Las Vegas,  NV,  1985.

16.    "Development and Evaluations of an  LC/MS/MS Protocol",  EPA/600/X-86/328,
      Dec. 1986.

17.    "An LC/MS Performance  Evaluation Study of Organophosphorus  Pesticides",
      EPA/600/X-89/006,  Jan.  1989.

18.    "A   Performance  Evaluation  Study  of   a   Liquid  Chromatography/Mass
      Spectrometry    Method    for    Tris-(2,3-Dibromopropyl)    Phosphate",
      EPA/600/X-89/135,  June 1989.

19.    "Liquid  Chromatography/Mass   Spectrometry   Performance   Evaluation  of
      Chlorinated Phenoxyacid Herbicides and Their Esters",  EPA/600/X-89/176,
      July 1989.

20.    "An Inter!aboratory Comparison  of an SW-846 Method for the Analysis of the
      Chlorinated Phenoxyacid Herbicides by LC/MS", EPA/600/X-90/133, June 1990.
                                   8321 -  22                        Revision 0
                                                                September 1994

-------
                                                                                                                           en
                                                                                                                         c o»
                                                                                                                         o •—i
CO


















0
1 — 1
1 — 1
o
0
o

o
:r
Q_
C£
0
h—
§
O
0£
X:
0
Q-
I

O
LU
O
"Z.
LiJ
s~
y
0
o
LlJ
Q—




























, 	 ^-^
03 CU C
!= E-r-
•i- -r- E
U_ H- -~*



CU
•— r— 0)
03 -r- OO
C -Q O3 . — «
*r"" ^j _£^ ^x
j [ T5T* fy ^— "^


c s-
O) 03
•i- CU ^~.
T3 C C
S- ^ E
CJ3 • — ' — *



r^Z
•I— < — s
+-> CU C
•r— E >r~
c: -r- E








O5 CU
•r- r- CU
-t-> ••— oo
*^~ f^ (& (^""N
C O _C 5?
H- 1 21 Q- 	 	












oo
OJ
^_)
>^
p—
03
C
«=C







LO LO

^
'o
^— ^^^
03 z
<~ C_J
-l-> CO
o cu o :r
O E O O






O LO
i — i







O 0



•~~^*
o

*-^ x:
i— O
O CU OJ O CU
LO +-> -C LO -t->
\^ ^j ^_j ^> 	 ^y
O 3 CU OS
LO 	 	 E LO >— -




00
3 --v
S- • •— i
O D1
-C • T3
a. cu cu
00 • — QC
O in
JC "O OO CU
a. c cu 01
03 >, 5-
c o o cu
O5 f~^ *~*
Dl E 0 to
S- O M -1-
o o <:o








if>

—
'o
c
05
_c
^_>
o cu
0 E
i — i — •






0
1— t







o
p
o

r5
_c
+^
cu
E
o cu
LO +J
^ "5
0 S
LO • — -


1
o
E CU
O 4-^
S- 03
t~l C"
-r- a.
t3 oo
i O
CO -C
" Q.
CVJ . — .
t^^'^—
1 >^
oo OL
•r- 0
S- S-
1— Q-









•X
o
rC
j^~
4_3
o cu
10 E

0 !S






LO
f-H







CM




jH"
o
c
03
-C
LO CU
CM E

LO > C
•r- X 3
S- O O
O C QL
r- 0) E
J= ^: o
(_j a. u
00 S-







0
t— i
*
'o
c
03
f
1 ^
LO CU
CM E

LO 
O3 a)
-= 0
•t-1 03
O CU
IO E E
^^^*^ 3

^~ — ' c
o
03
5:
i— i
o
•
0

II

^£

cu
S-
cu
-SZ
3


*
•r- CU
> .a
cu E
Q£ CU
CL
CU
OO



















CO
CM

'
r— 1
CM
CO
00


































-------
                                   TABLE 2.
             COMPOUNDS AMENABLE TO THERMOSPRAY MASS SPECTROMETRY
      Disperse Azo Dyes       Alkaloids
      Methine Dyes            Aromatic ureas
      Arylmethane Dyes        Amides
      Coumarin Dyes           Amines
      Anthraquinone Dyes      Amino acids
      Xanthene Dyes           Organophosphorus Compounds
      Flame retardants        Chlorinated Phenoxyacid Compounds
                                   TABLE 3.
              LIMITS OF DETECTION (LOD) AND METHOD SENSITIVITIES
                        FOR DISPERSE RED  1  AND CAFFEINE
Compound

Disperse Red 1


Caffeine


Mode

SRM
Single Quad
CAD
SRM
Single Quad
CAD
LOD
(P9)
180
600
2,000
45
84
240
EQL(7s)
(P9)
420
1400
4700
115
200
560
EQL(lOs)
(pg)
600
2000
6700
150
280
800
EQL = Estimated Quantitation Limit

Data from Reference 16.
                                   8321  -  24                         Revision 0
                                                                September 1994

-------
                                   TABLE 4.
            PRECISION  AND ACCURACY  COMPARISONS  OF MS AND  MS/MS  WITH
       HPLC/UV  FOR ORGANIC-FREE  REAGENT  WATER SPIKED WITH DISPERSE RED  1

Sample
Spike 1
Spike 2
RPD

HPLC/UV
82.2 ± 0.2
87.4 ± 0.6
6.1%
Percent
MS
92.5 ± 3.7
90.2 ± 4.7
2.5%
Recovery
CAD
87.6 ± 4.6
90.4 ± 9.9
3 . 2%

SRM
95.5 ± 17.1
90.0 ± 5.9
5.9%
Data from Reference 16.
                                   TABLE 5.
           PRECISION AND ACCURACY COMPARISONS OF MS AND MS/MS WITH
          HPLC/UV  FOR  MUNICIPAL  WASTEWATER  SPIKED WITH  DISPERSE  RED  1
Percent Recovery
Sample
Spike 1
Spike 2
RPD
HPLC/UV
93.4 ± 0.3
96.2 + 0.1
3.0%
MS
102.0 + 31
79.7 ± 15
25%
CAD
82.7 + 13
83.7 + 5.2
1 . 2%
Data from Reference 16.
                                  8321  - 25
    Revision 0
September 1994

-------
                                   TABLE 6.
         RESULTS FROM ANALYSES OF ACTIVATED SLUDGE PROCESS WASTEWATER
Sample
5 mg/L Spiking
Concentration
1
1-D
2
3
RPD
Unspiked
Sample
1
1-D
2
3
RPD
Recovery
HPLC/UV
0.721 + 0.003
0.731 + 0.021
0.279 + 0.000
0.482 ± 0.001
1.3%

0.000
0.000
0.000
0.000
--
of Disperse Red 1
MS
0.664 + 0.030
0.600 + 0.068
0.253 + 0.052
0.449 + 0.016
10.1%

0.005 ± 0.0007
0.006 ± 0.001
0.002 ± 0.0003
0.003 ± 0.0004
18.2%
(mq/L)
CAD
0.796 + 0.008
0.768 + 0.093
0.301 ± 0.042
0.510 + 0.091
3 . 6%

<0.001
<0.001
<0.001
<0.001
--
Data from Reference 16.
                                   8321  -  26                         Revision 0
                                                                September 1994

-------
                  TABLE 7.
CALIBRATION MASSES AND % RELATIVE ABUNDANCES
                  OF PEG 400
Mass
18.0
35.06
36.04
50.06
77.04
168.12
212.14
256.17
300.20
344.22
388.25
432.28
476.30
520.33
564.35
608.38
652.41
653.41
696.43
697.44
% Relative
Abundances8
32.3
13.5
40.5
94.6
27.0
5.4
10.3
17.6
27.0
45.9
64.9
100
94.6
81.1
67.6
32.4
16.2
4.1
8.1
2.7
   Intensity is normalized to mass 432,
                  8321  -  27                         Revision 0
                                               September 1994

-------
                  TABLE 8.
CALIBRATION MASSES AND % RELATIVE ABUNDANCES
                 OF PEG 600
Mass
18.0
36.04
50.06
77.04
168.12
212.14
256.17
300.20
344.22
388.25
432.28
476.30
520.33
564.35
608.38
652.41
653.41
696.43
% Relative
Abundances8
4.7
11.4
64.9
17.5
9.3
43.9
56.1
22.8
28.1
38.6
54.4
64.9
86.0
100
63.2
17.5
5.6
1.8
         Intensity  is  normalized  to  mass  564.
                  8321  -  28                         Revision 0
                                               September 1994

-------
                                   TABLE 9.
                 RETENTION TIMES AND THERMOSPRAY MASS SPECTRA
                         OF  ORGANOPHOSPHORUS  COMPOUNDS
Compound
Monocrotophos
Trichlorfon
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
Retention Time
(minutes)
1:09
1:22
1:28
4:40
9:16
9:52
10:52
13:30
13:55
18:51
Mass Spectra
(% Relative Abundance)8
241 (100), 224 (14)
274 (100), 257 (19), 238 (19)
230 (100), 247 (20)
238 (100), 221 (40)
398 (100), 381 (23), 238 (5),
221 (2)
326 (10), 309 (100)
281 (100), 264 (8), 251 (21),
234 (48)
278 (4), 261 (100)
292 (10), 275 (100)
315 (100), 299 (15)
  a  For molecules containing Cl,  Br and S, only the base peak of the isotopic
  cluster is listed.

Data from Reference 17.
                                  8321  - 29
    Revision 0
September 1994

-------
                                   TABLE  10.
               PRECISION AND METHOD DETECTION LIMITS (MDLs) FOR
                      ORGANOPHOSPHORUS  COMPOUND  STANDARDS
Compound
Dichlorvos
Dimethoate
Phorate
Disulfoton
Fensulfothion
Naled
Merphos
Methyl
parathion
Ion
238
230
261
275
309
398
299
281
Standard
Quantitation
Concentration
(ng/juL)
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
2
12.5
25
50
%RSD
16
13
5.7
4.2
2.2
4.2
13
7.3
0.84
14
7.1
4.0
2.2
14
6.7
3.0
4.1
9.2
9.8
2.5
9.5
9.6
5.2
6.3
5.5
17
3.9
5.3
7.1
4.8
1.5
MDL (ng)
4
2
2
1
0.4
0.2
1
30
Data from Reference 17.
                                   8321  -  30
    Revision 0
September 1994

-------
                                   TABLE 11.
     SINGLE OPERATOR ACCURACY AND PRECISION FOR LOW CONCENTRATION DRINKING
     WATER  (A), LOW CONCENTRATION SOIL  (B), MEDIUM CONCENTRATION DRINKING
                 WATER  (C),  MEDIUM CONCENTRATION SEDIMENT  (D)
Average
Recovery
Compound (%)
A
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
B
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
C
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos
D
Dimethoate
Dichlorvos
Naled
Fensulfothion
Methyl parathion
Phorate
Disulfoton
Merphos

70
40
0.5
112
50
16
3.5
237

16
ND
ND
45
ND
78
36
118

52
146
4
65
85
10
2
101

74
166
ND
72
84
58
56
78
Standard
Deviation

7.7
12
1.0
3.3
28
35
8
25

4


5
--
15
7
19

4
29
3
7
24
15
1
13

8.5
25
--
8.6
9
6
5
4
Spike
Amount
ug/L
5
5
5
5
10
5
5
5
uq/kq
50
50
50
50
100
50
50
50
uq/L
50
50
50
50
100
50
50
50
mg/kq
2
2
2
2
3
2
2
2
Range of
Recovery
(%)

54
14
0
106
0
0
0
187

7
-
-
34
-
48
22
81

43
89
0
51
37
0
0
75

57
115
-
55
66
46
47
70

- 85
- 64
- 2
- 119
- 105
- 86
- 19
-287

- 24
-
-
- 56
-
- 109
- 49
- 155

- 61
- 204
- 9
- 79
- 133
- 41
- 4
- 126

- 91
- 216
-
- 90
- 102
- 70
- 66
- 86
Number
of
Analyses

15
15
15
15
15
15
15
15

15
15
15
15
15
15
15
15

12
12
12
12
12
12
12
12

15
15
15
15
15
15
15
12
Data from Reference 17.
                                  8321  - 31
    Revision 0
September 1994

-------
                                   TABLE 12
          SINGLE OPERATOR ACCURACY AND PRECISION FOR MUNICIPAL WASTE
           WATER (A), DRINKING WATER (B),  CHEMICAL SLUDGE WASTE (C)
Average
Recovery
Compound (%)
Tris-BP (A) 25
(B) 40
(C) 63
Spike Range
Standard Amount of % Number of
Deviation (ng/juL) Recovery Analyses
8.0 2 41 - 9.0 15
5.0 2 50-30 12
11 100 84-42 8
Data from Reference 18.
                                   8321  -  32
    Revision 0
September 1994

-------
              fc
              Q.UJ
CQ
cr:
o
CO

             oo  >
             •X   CU
             O Q
              c/>  a)
              *  Q

+j O)
 Q
               OJ
 S- fO
 CU d)
 > 5-
<£.  S-
                                                                                                .,- CO
                                                                                                 > -Q
                                                                                                 CU E
                                                                                                O£ CU
                                                                                                    +-i
                                                                                                    a.
                                                                                                    cu
                                                                                                    oo
                              ro
                              CO
                 ro
                 CVJ
                 oo
                 i—
                 10
                 i--.
                 *3-
                 LO
                                                                                                                               ro
                                                                                                                               ro
                                                                                                                               CM
                                                                                                                               CO
                                                                                                                               00
                 oo
                 CNJ
                 CM 00  O O
                 oo LO  cr> ro
                 t^ in  C5 o
                        CVJ CVJ
If) r-t ^- CT>
r~- en i~^ i—
10 o 10 ro
CM in i— oo
                 o o o o
                 ID CD LO o
                    i—I r—I CM
                                        00
                                         CU
                                         o
                                         c
                                         CU
                                         s_
                                         CU
                                        It-
                                         CD
                                                       o
                                                       S-
                                                       (O
                                                      O

-------
                                   TABLE 14
       LIMITS OF DETECTION (LOD) IN THE POSITIVE AND NEGATIVE ION MODES
          FOR THE CHLORINATED PHENOXYACID HERBICIDES AND FOUR ESTERS
Compound
Dalapon
Dicamba
2,4-D
MCPA
Dichlorprop
MCPP
2,4,5-T
2,4,5-TP (Silvex)
Dinoseb
2,4-DB
2,4-D,Butoxy
ethanol ester
2,4,5-T,Butoxy
ethanol ester
2,4,5-T, Butyl
ester
2,4-D,ethyl-
hexyl ester
Positive Mode
Quantitation
Ion
Not detected
238 (M+NH4)+
238 (M+NH4)+
218 (M+NH4)+
252 (M+NH4}+
232 (M+NHJ +
272 (M+NH4)+
286 (M+NH4)+
228 (M+NH4-NO)+
266 (M+NH4)+
321 (M+H)+

372 (M+NH4)+

328 (M+NH4)+

350 (M+NHJ +

LOD
(ng)

13
2.9
120
2.7
5.0
170
160
24
3.4
1.4

0.6

8.6

1.2

Negative Mode
Quantitation
Ion
141 (M-H)'
184 (M-HC1)'
184 (M-HC1)-
199 (M-l)-
235 (M-l)-
213 (M-l)-
218 (M-HC1)'
269 (M-l)'
240 (M)-
247 (M-l)-
185 (M-CeH^OJ-

195 (M-C8H1503)-

195 (M-C6Hn02)-

161 (M-C10H1903)-

LOD
(ng)
11
3.0
50
28
25
12
6.5
43
19
110








Data from Reference 19.
                                   8321  -  34
    Revision 0
September 1994

-------
                                   TABLE  15
               SINGLE LABORATORY OPERATOR ACCURACY AND PRECISION
                  FOR THE CHLORINATED  PHENOXYACID HERBICIDES
Compound

Dicamba
2,4-D
MCPA
MCPP
Dichlorprop
2,4,5-T
Si 1 vex
2,4-DB
Dinoseb
Dalapon
2,4-D, ester

Dicamba
2,4-D
MCPA
MCPP
Dichlorprop
2,4,5-T
Silvex
2,4-DB
Dinoseb
Dalapon
2,4-D, ester

Dicamba
2,4-D
MCPA
MCPP
Dichlorprop
2,4,5-T
Silvex
2,4-DB
Dinoseb
Dalapon
2,4-D, ester
(a)
Average Standard
Recovery(%) Deviation
LOW LEVEL
63
26
60
78
43
72
62
29
73
ND
73
HIGH LEVEL
54
60
67
66
66
61
74
83
91
43
97
LOW
117
147
167
142
ND
134
121
199
76
ND
180
DRINKING WATER
22
13
23
21
18
31
14
24
11
ND
17
DRINKING WATER
30
35
41
33
33
23
35
25
10
9.6
19
LEVEL SAND
26
23
79
39
ND
27
23
86
74
ND
58
Spike
Amount
M9/L
5
5
5
5
5
5
5
5
5
5
5
M9/L
50
50
50
50
50
50
50
50
50
50
50
M9/9
.1
.1
.1
.1
.1
.1
.1
.1
.1
.1
.1
Range of
Recovery
(%)

33
0
37
54
0
43
46
0
49

48

26
35
32
35
27
44
45
52
76
31
76

82
118
78
81

99
85
0
6

59

- 86
- 37
- 92
- 116
- 61
- 138
- 88
- 62
- 85
ND
- 104

- 103
- 119
- 128
- 122
- 116
- 99
- 132
- 120
- 102
- 56
- 130

- 147
- 180
- 280
- 192
ND
- 171
- 154
- 245
- 210
ND
- 239
Number
of
Analyses

9
9
9
9
9
9
9
9
9
9
9

9
9
9
9
9
9
9
9
9
6
9

10
10
10
10
10
10
10
10
10
10
7
la)All  recoveries  are  in  negative  ionization  mode,  except  for 2,4-D,ester.
ND = Not Detected.
                                   8321  -  35
    Revision 0
September 1994

-------
                               TABLE 15 (cont.)
               SINGLE  LABORATORY  OPERATOR  ACCURACY'AND  PRECISION
                  FOR THE CHLORINATED PHENOXYACID HERBICIDES


Compound
(a)
Average
Recovery(%)

Standard
Deviation

Spike
Amount
Range of
Recovery
(%)
Number
of
Analyses
Dicamba
2,4-D
MCPA
MCPP
Dichlorprop
2,4,5-T
Si 1 vex
2,4-DB
Dinoseb
Dalapon
2,4-D,ester
Dicamba
2,4-D
MCPA
MCPP
Dichlorprop
2,4,5-T
Si 1 vex
2,4-DB
Dinoseb
Dalapon
2,4-D,ester
Dicamba
2,4-D
MCPA
MCPP
Dichlorprop
2,4,5-T
Si 1 vex
2,4-DB
Dinoseb
Dalapon
2,4-D,ester
                     HIGH LEVEL SAND
153
218
143
158
 92
160
176
145
114
287
 20
33
27
30
34
37
29
34
22
28
86
3.6
                 LOW LEVEL MUNICIPAL ASH
83
IND
ND
ND
ND
27
68
ND
44
ND
29
22
ND
ND
ND
ND
25
38
ND
13
ND
23
 66
8.7
3.2
 10
 ND
2.9
6.0
 ND
 16
 ND
1.9
 21
4.8
4.8
4.3
 ND
1.2
3.1
 ND
6.8
 ND
1.7
M9/9

  1
  1
  1
  1
  1
  1
  1
  1
  1
  1
  1
  .1
  .1
  .1
  .1
  .1
  .1
  .1
  .1
  .1
  .1
  .1
                 HIGH LEVEL MUNICIPAL ASH   M9/9
119
187
111
115
51
131
141
110
65
166
17
- 209
- 276
- 205
- 226
- 161
- 204
- 225
- 192
- 140
- 418
- 25
48 - 104
   ND
   ND
   ND
   ND
 0 - 60
22 - 128
   ND
26 - 65
   ND
 0 - 53
          41  -  96
            5  -  21
            0  -  10
          4.7  -  16
              ND
            0  -  3.6
            2.8  - 12
              ND
            0  -  23
              ND
            0  -  6.7
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 7
9
9
9
9
9
9
9
9
9
9
6
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 9
                 6
lalAll  recoveries are in negative ionization  mode,  except  for 2,4-D,ester.
ND = Not Detected.
                                   8321 - 36
                                                 Revision 0
                                             September 1994

-------
                                   TABLE 16
                  MULTI LABORATORY ACCURACY  AND PRECISION DATA
                  FOR THE CHLORINATED PHENOXYACID HERBICIDES
        Compounds
Spiking
Concentration
    Mean
(% Recovery)8
Data from Reference 20.
a  Mean  of duplicate data from 3 laboratories.
b  % RSD of duplicate data from 3 laboratories.
   %  Relative
Standard Deviation6
500 mg/L
2,4,5-T
2,4,5-T,butoxy
2,4-D
2,4-DB
Dalapon
Dicamba
Dichlorprop
Dinoseb
MCPA
MCPP
Si 1 vex
50 mg/L
2,4,5-T
2,4,5-T,butoxy
2,4-D
2,4-DB
Dalapon
Dicamba
Dichlorprop
Dinoseb
MCPA
MCPP
Silvex
5 mq/L
2,4,5-T
2,4,5-T,butoxy
2,4-D
2,4-DB
Dalapon
Dicamba
Dichlorprop
Dinoseb
MCPA
MCPP
Silvex

90
90
86
95
83
77
84
78
89
86
96

62
85
64
104
121
90
96
86
96
76
65

90
99
103
96
150
105
102
108
94
98
87

23
29
17
22
13
25
20
15
11
12
27

68
9
80
28
99
23
15
57
20
74
71

28
17
31
21
4
12
22
30
18
15
15
                                   8321  -  37
                                           Revision  0
                                       September  1994

-------
                                   TABLE 17
           COMPARISON OF LODs: METHOD 8151 vs. METHOD 8321
Compound
Method 8151
 LOD(Mg/L)
Method 8321
LOD (Mg/L)
lonization
   Mode
Dalapon
Dicamba
2,4-D
MCPA
Dichloroprop
MCPP
2,4,5-T
2,4,5-TP (Silvex)
2,4-DB
Dinoseb
     1.3
     0.8
     0.2
     0.06
     0.26
     0.09
     0.08
     0.17
     0.8
     0.19
  1.1
  0.3
  0.29
  2.8
  0.27
  0.50
  0.65
  4.3
  0.34
  1.9
                                   8321 - 38
                                       Revision 0
                                   September 1994

-------
                        FIGURE 1.
    SCHEMATIC OF THE THERMOSPRAY PROBE AND ION SOURCE
 r
&
                      Flange
       Source
      Mounting
                   Ion Sampling
                      Cone
                    Ions
Electron  Vaporizer
 Beam  .x Probe
             T,
\
                                                                   — 1C
i
                   Vapor    ||  Heater  Vaporizer
                Temperature |         Coupling
                 T4        Block
                       Temperature
                        T.
                        8321 - 39
                                                          Revision  0
                                                      September  1994

-------
            FIGURE 2.
THERMOSPRAY SOURCE WITH WIRE-REPELLER
   (High sensitivity configuration)
    CERAMIC INSULATOR
    WIRE REPELLER
              8321 - 40
   Revision 0
September 1994

-------
            FIGURE 3.
THERMOSPRAY SOURCE WITH WIRE-REPELLER
        (CAD configuration)
  CERAMIC INSULATOR
  WIRE REPELLER
            8321  - 41
   Revision 0
September 1994

-------
                            METHOD 8321
           SOLVENT EXTRACTABLE NON-VOLATILE COMPOUNDS BY
HIGH PERFORMANCE LIQUID CHROMATOGRAPHY/THERMOSPRAY/MASS SPECTROMETRY
             (HPLC/TSP/MS) OR  ULTRAVIOLET  (UV) DETECTION
                              8321  -  42
    Revision 0
September 1994

-------
                                  METHOD  8330

                     NITROAROMATICS  AND NITRAMINES  BY  HIGH
                   PERFORMANCE LIQUID CHROMAT06RAPHY (HPLC)
1.0   SCOPE AND APPLICATION

      1.1   Method 8330 is intended for the trace analysis of explosives residues
by high performance liquid chromatography using a UV detector.  This method is
used to determine the concentration of the following compounds  in a water, soil,
or sediment matrix:
Compound
Abbreviation
CAS No8
Octahydro-l,3,5,7-tetranitro-l,3,5,7-tetrazocine       HMX
Hexahydro-l,3,5-trinitro-l,3,5-triazine                RDX
1,3,5-Trinitrobenzene                                  1,3,5-TNB
1,3-Dinitrobenzene                                     1,3-DNB
Methyl-2,4,6-tri ni trophenylni trami ne                   Tetryl
Nitrobenzene                                           NB
2,4,6-Trinitrotoluene                                  2,4,6-TNT
4-Amino-2,6-dinitrotoluene                             4-Am-DNT
2-Amino-4, 6-dinitrotoluene                            2-Am-DNT
2,4-Dinitrotoluene                                     2,4-DNT
2,6-Dinitrotoluene                                     2,6-DNT
2-Nitrotoluene                                         2-NT
3-Nitrotoluene                                         3-NT
4-Nitrotoluene                                         4-NT
                  2691-41-0
                   121-82-4
                    99-35-4
                    99-65-0
                   479-45-8
                    98-95-3
                   118-96-7
                  1946-51-0
                355-72-78-2
                   121-14-2
                   606-20-2
                    88-72-2
                    99-08-1
                    99-99-0
a  Chemical Abstracts Service Registry number

      1.2   Method 8330  provides a  salting-out  extraction procedure  for  low
concentration (parts  per trillion, or nanograms per liter)  of explosives residues
in surface  or  ground water.   Direct injection  of diluted and  filtered water
samples can be used for water samples of higher concentration (See Table 1).

      1.3   All  of these  compounds  are  either  used  in  the manufacture  of
explosives or are the degradation products of compounds used for that purpose.
When making stock solutions for calibration, treat each explosive compound with
caution.  See NOTE in Sec. 5.3.1 and Sec.  11 on Safety.

      1.4   The  estimated  quantitation  limits  (EQLs)   of  target  analytes
determined by Method 8330 in water and soil  are presented in Table 1.

      1.5   This method  is restricted to use by  or under the  supervision  of
analysts  experienced  in  the  use of  HPLC,  skilled  in the  interpretation  of
chromatograms, and experienced in handling explosive materials.  (See Sec. 11.0
                                   8330 - 1
                 Revision 0
            September  1994

-------
on SAFETY.)  Each  analyst must  demonstrate  the  ability to generate acceptable
results with this method.
2.0   SUMMARY OF METHOD

      2.1   Method 8330 provides high  performance liquid chromatographic (HPLC)
conditions for  the  detection  of ppb levels of certain  explosives  residues in
water, soil  and  sediment matrix.  Prior to use of this method, appropriate sample
preparation techniques must be used.

      2.2   Low-Level Salting-out Method With No Evaporation:   Aqueous samples
of low concentration  are  extracted  by a salting-out  extraction procedure with
acetonitrile and sodium chloride.  The small volume  of  acetonitrile that remains
undissolved  above  the salt water  is  drawn off  and  transferred to  a smaller
volumetric flask.   It is back-extracted by vigorous  stirring  with  a specific
volume of salt water.  After equilibration, the phases are allowed to separate
and  the  small  volume  of acetonitrile  residing  in  the  narrow  neck of  the
volumetric flask is removed using a  Pasteur pi pet.  The concentrated extract is
diluted 1:1 with reagent grade water.  An aliquot is separated on a C-18 reverse
phase column, determined at 254 nm,  and confirmed on a CN reverse phase column.

      2.3   High-level  Direct  Injection Method:   Aqueous  samples  of  higher
concentration can be diluted 1/1  (v/v)  with methanol or acetonitrile, filtered,
separated on a C-18 reverse phase column, determine at 254 nm,  and confirmed on
a CN reverse phase column.  If HMX  is  an important target analyte, methanol is
preferred.

      2.4   Soil and  sediment  samples are  extracted  using  acetonitrile  in an
ultrasonic bath, filtered and chromatographed as in Sec. 2.3.


3.0   INTERFERENCES

      3.1   Solvents, reagents, glassware  and other sample processing hardware
may yield discrete artifacts and/or elevated basel ines,  causing misinterpretation
of the chromatograms.   All  of  these materials must be demonstrated to be free
from interferences.

      3.2   2,4-DNT and 2,6-DNT elute  at similar retention  times (retention time
difference of 0.2 minutes).  A large  concentration of one isomer may mask the
response of  the other  isomer.   If it  is  not apparent that both  isomers are
present  (or are not detected), an isomeric  mixture should be reported.

      3.3   Tetryl  decomposes  rapidly in methanol/water solutions,  as well as
with heat.  All  aqueous samples expected to  contain tetryl should be diluted with
acetonitrile prior  to filtration and acidified to pH <3.  All   samples expected
to contain tetryl  should not be exposed to temperatures above room temperature.

      3.4   Degradation products of tetryl  appear as a shoulder  on the 2,4,6-TNT
peak.  Peak heights rather than peak areas  should be used when tetryl  is present
in  concentrations  that  are   significant   relative  to the concentration  of
2,4,6-TNT.

                                    8330  -  2                        Revision  0
                                                                September 1994

-------
4.0   APPARATUS AND MATERIALS
      4.1   HPLC system
            4.1.1 HPLC - equipped with a pump capable of achieving 4000 psi, a
      100 /zl loop injector and a 254 nm UV  detector   (Perkin Elmer Series 3, or
      equivalent).   For the  low concentration option,  the detector  must be
      capable of a stable baseline at 0.001 absorbance units full scale.
            4.1.2 Recommended Columns:
                  4.1.2.1     Primary column:  C-18 Reverse phase HPLC column,
            25 cm x 4.6 mm (5 jum),  (Supelco LC-18, or equivalent).
                  4.1.2.2     Secondary column:  CN Reverse phase HPLC column,
            25 cm x 4.6 mm (5 /^m),  (Supelco LC-CN, or equivalent).
            4.1.3 Strip chart recorder.
            4.1.4 Digital integrator (optional).
            4.1.5 Autosampler (optional).
      4.2   Other Equipment
            4.2.1 Temperature controlled ultrasonic bath.
            4.2.2 Vortex mixer.
            4.2.3 Balance, + 0.0001 g.
            4.2.4 Magnetic stirrer with stirring pellets.
            4.2.5 Water bath  -  Heated,  with  concentric  ring cover,  capable of
      temperature control (± 5°C).  The bath should be used in a hood.
            4.2.6 Oven - Forced air, without heating.
      4.3   Materials
            4.3.1 High pressure  injection  syringe - 500 p,L,  (Hamilton liquid
      syringe or equivalent).
            4.3.2 Disposable cartridge filters - 0.45 /im Teflon  filter.
            4.3.3 Pipets - Class A, glass,  Appropriate sizes.
            4.3.4 Pasteur pipets.
            4.3.5 Scintillation Vials - 20 mL, glass.
            4.3.6 Vials - 15 mL, glass, Teflon-lined cap.
                                   8330 - 3                         Revision 0
                                                                September 1994

-------
            4.3.7 Vials- 40 ml,  glass,  Teflon-lined cap.

            4.3.8 Disposable syringes - Plastipak, 3 ml and  10 ml or equivalent.

            4.3.9 Volumetric  flasks  -  Appropriate  sizes  with  ground  glass
      stoppers,  Class A.

            NOTE: The 100 ml and 1 L volumetric flasks  used  for magnetic stirrer
                  extraction must be round.

            4.3.10      Vacuum desiccator -  Glass.

            4.3.11      Mortar and pestle -  Steel.

            4.3.12      Sieve -  30 mesh.

            4.3.13      Graduated cylinders  - Appropriate sizes.

      4.4   Preparation of Materials

            4.4.1 Prepare all materials to be used as described  in Chapter 4 for
      semivolatile organics.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall  be  used  in all tests.  Unless
otherwise  indicated,  it is  intended  that all  reagents  shall  conform  to  the
specifications of the Committee  on Analytical  Reagents of the American Chemical
Society, where such  specifications  are  available.   Other  grades may be used,
provided it is first,  ascertained that the reagent is of sufficiently high purity
to permit its use without lowering the accuracy of the determination.

            5.1.1 Acetonitrile,  CH3CN  -  HPLC grade.

            5.1.2 Methanol, CH3OH -  HPLC  grade.

            5.1.3 Calcium chloride, CaCl2 -  Reagent grade.   Prepare an aqueous
      solution of 5 g/L.

            5.1.4 Sodium  chloride,  NaCl, shipped  in  glass bottles  -  reagent
      grade.

      5.2   Organic-free reagent water  -  All  references to  water in this method
refer to organic-free reagent water, as defined in Chapter One.

      5.3   Stock Standard  Solutions

            5.3.1 Dry each solid analyte  standard to constant weight in a vacuum
      desiccator  in  the  dark.   Place about  0.100 g (weighed  to 0.0001  g) of a
      single  analyte  into  a 100  ml  volumetric flask and  dilute to volume with
      acetonitrile.   Invert  flask  several  times  until  dissolved.   Store in


                                   8330  - 4                         Revision 0
                                                                September 1994

-------
refrigerator at 4°C in the dark.  Calculate  the concentration of the stock
solution from the  actual weight used (nominal  concentration = 1,000 mg/L).
Stock solutions may be used for up to one year.

      NOTE: The HMX,  RDX,  Tetryl,  and 2,4,6-TNT are  explosives  and the
            neat material  should be handled carefully.  See SAFETY in Sec.
            11 for guidance.   HMX, RDX,   and Tetryl  reference materials
            are  shipped  under  water.   Drying  at  ambient  temperature
            requires several days.  DO NOT DRY AT HEATED TEMPERATURES!

5.4   Intermediate Standards Solutions

      5.4.1 If both 2,4-DNT and 2,6-DNT are to be determined, prepare two
separate intermediate stock solutions  containing  (1)  HMX, RDX, 1,3,5-TNB,
1,3-DNB, NB, 2,4,6-TNT, and 2,4-DNT and (2) Tetryl, 2,6-DNT, 2-NT,  3-NT,
and 4-NT.   Intermediate  stock standard solutions  should be prepared at
1,000 M9/L, in acetonitrile when analyzing soil  samples, and in methanol
when analyzing aqueous samples.

      5.4.2 Dilute the two concentrated intermediate stock solutions, with
the appropriate solvent,  to prepare intermediate standard solutions that
cover  the  range  of  2.5  -  1,000  jug/L.    These   solutions  should  be
refrigerated on preparation, and may be used for 30 days.

      5.4.3 For the low-level method, the analyst must conduct a detection
limit study and devise dilution series appropriate to the desired range.
Standards for the low level method must be prepared immediately prior to
use.

5.5   Working standards

      5.5.1 Calibration  standards  at  a  minimum  of  five  concentration
levels should be prepared through dilution of the intermediate standards
solutions by 50% (v/v) with 5 g/L calcium chloride solution  (Sec. 5.1.3).
These solutions must be refrigerated  and  stored  in the dark, and prepared
fresh on the day of calibration.

5.6   Surrogate Spiking Solution

      5.6.1 The analyst should monitor the performance of the extraction
and  analytical  system as  well as  the  effectiveness  of  the method in
dealing with  each sample  matrix  by  spiking  each  sample,   standard and
reagent water blank  with  one  or two  surrogates  (e.g.,   analytes  not
expected to be present in the sample).

5.7   Matrix Spiking Solutions

      5.7.1 Prepare matrix  spiking  solutions in methanol such  that the
concentration in the  sample  is five times  the  Estimated Quantitation Limit
(Table 1).  All target analytes should be included.
                             8330 - 5                         Revision 0
                                                          September 1994

-------
      5.8   HPLC Mobile Phase

            5.8.1 To prepare 1 liter of mobile phase, add 500 ml of methanol to
      500 ml of organic-free reagent water.


6.0   SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1   Follow  conventional  sampling  and  sample  handling procedures  as
specified for semivolatile organics in Chapter Four.

      6.2   Samples and  sample extracts must  be stored in  the dark  at  4"C.
Holding times are the same as for semivolatile organics.


7.0   PROCEDURE

      7.1   Sample Preparation

            7.1.1 Aqueous Samples:   It is highly  recommended that process waste
      samples be  screened  with the high-level  method to determine  if the low
      level method (1-50 /ig/L) is required.  Most groundwater samples will fall
      into the low level method.

                  7.1.1.1     Low-Level Method (salting-out extraction)

                        7.1.1.1.1   Add 251.3  g  of  sodium  chloride to a  1  L
                  volumetric  flask (round).    Measure  out 770  mL  of  a water
                  sample (using a 1 L graduated cylinder) and  transfer it to the
                  volumetric flask containing  the salt.  Add  a stir  bar  and mix
                  the contents at maximum  speed on  a magnetic stirrer until the
                  salt is completely dissolved.

                        7.1.1.1.2   Add 164 mL  of acetonitrile  (measured with a
                  250 mL graduated  cylinder) while the solution  is being  stirred
                  and stir for an additional 15 minutes.  Turn off  the  stirrer
                  and allow the phases to separate for 10 minutes.

                        7.1.1.1.3   Remove the acetonitrile (upper) layer (about
                  8 mL)  with  a Pasteur pipet   and  transfer it  to a  100 mL
                  volumetric flask (round). Add  10 mL of fresh acetonitrile to
                  the water sample in the 1 L flask.  Again  stir the  contents of
                  the flask  for 15 minutes followed by 10  minutes for phase
                  separation.  Combine the  second acetonitrile portion with the
                  initial extract.  The inclusion of a few drops of salt water
                  at this point is unimportant.

                        7.1.1.1.4   Add 84 mL of  salt water (325 g  Nad per  1000
                  ml. of reagent water)  to the acetonitrile extract  in the 100 mL
                  volumetric flask.  Add a stir  bar and stir the contents on a
                  magnetic stirrer  for 15  minutes,  followed  by 10 minutes for
                  phase separation.  Carefully transfer the acetonitrile phase


                                   8330 - 6                         Revision 0
                                                                September  1994

-------
      to a 10 ml graduated  cylinder using a Pasteur pipet.  At this
      stage, the amount of water transferred with the acetonitrile
      must be minimized.  The water contains  a high concentration of
      NaCl  that produces  a  large peak  at  the  beginning  of  the
      chromatogram,  where   it   could  interfere  with   the  HMX
      determination.

            7.1.1.1.5   Add an additional 1.0 ml of acetonitrile to
      the 100 ml volumetric flask.  Again stir the contents of the
      flask  for 15  minutes,  followed  by  10 minutes  for  phase
      separation.  Combine  the second  acetonitrile portion with the
      initial extract in the 10 ml graduated cylinder (transfer to
      a  25  ml  graduated  cylinder if the   volume  exceeds  5 ml).
      Record the total volume of acetonitrile extract  to the nearest
      0.1 ml.  (Use this as the volume of total extract  [Vt] in the
      calculation of concentration after converting  to juL).   The
      resulting extract, about 5  - 6  ml, is then  diluted 1:1 with
      organic-free  reagent  water (with  pH  <3   if  tetryl  is  a
      suspected analyte) prior to analysis.

            7.1.1.1.6   If the diluted extract is turbid, filter it
      through a 0.45 - jum Teflon  filter using a disposable syringe.
      Discard the first  0.5 ml of filtrate, and retain the remainder
      in a Teflon-capped vial for RP-HPLC analysis as in Sec. 7.4.

      7.1.1.2     High-level Method

            7.1.1.2.1   Sample filtration:   Place  a 5 ml aliquot of
      each  water sample  in  a  scintillation  vial,  add  5  ml  of
      acetonitrile, shake thoroughly,  and filter  through a 0.45-jum
      Teflon filter using a disposable syringe.   Discard the first
      3 ml of filtrate,  and retain the remainder in a Teflon-capped
      vial for  RP-HPLC  analysis  as in Sec.  7.4.   HMX quantitation
      can  be  improved   with  the use  of  methanol   rather  than
      acetonitrile for dilution before filtration.

7.1.2 Soil and Sediment Samples

      7.1.2.1     Sample homogenization:  Dry soil  samples in  air at
room temperature or colder to a constant weight,  being careful  not
to expose the samples to direct sunlight.  Grind and homogenize the
dried sample thoroughly in an acetonitrile-rinsed mortar to pass a
30 mesh sieve.

      NOTE:  Soil samples should be screened  by Method 8515 prior to
            grinding in  a mortar  and pestle  (See Safety Sec. 11.2).

      7.1.2.2     Sample extraction

            7.1.2.2.1   Place a 2.0 g  subsample of each soil sample
      in a 15 ml glass vial.   Add  10.0 ml of acetonitrile, cap with
                       8330 - 7                         Revision 0
                                                    September 1994

-------
            Teflon-lined cap,  vortex swirl for one minute, and place in a
            cooled ultrasonic bath for 18 hours.

                  7.1.2.2.2   After sonication, allow sample to settle for
            30 minutes.  Remove 5.0  ml of supernatant,  and combine with
            5.0 ml of calcium chloride solution  (Sec. 5.1.3)  in  a 20 ml
            vial.  Shake, and let stand for  15 minutes.

                  7.1.2.2.3   Place  supernatant  in  a  disposable  syringe
            and filter through a 0.45-jum Teflon filter.   Discard first 3
            mL and retain remainder  in a Teflon-capped  vial  for RP-HPLC
            analysis as in Sec.  7.4.

7.2   Chromatographic Conditions (Recommended)

      Primary Column:   C-18 reverse phase  HPLC column,  25-cm
                        x 4.6-mm,  5 /urn, (Supelco LC-18 or equivalent).

      Secondary Column: CN reverse phase HPLC column,  25-cm x
                        4.6-mm,  5 /j,m, (Supelco LC-CN or
                        equivalent).

      Mobile Phase:     50/50 (v/v) methanol/organic-free
                        reagent water.

      Flow Rate:        1.5 mL/min

      Injection volume: 100-^L

      UV Detector:      254 nm

7.3   Calibration of HPLC

      7.3.1 All electronic equipment is allowed to warm up for 30 minutes.
During this period,  at  least  15 void volumes of  mobile  phase are passed
through the  column  (approximately 20 min  at 1.5 mL/min)  and continued
until the baseline is level at the UV detector's greatest sensitivity.

      7.3.2 Initial Calibration.   Injections  of each calibration standard
over the concentration  range  of  interest are made sequentially into the
HPLC in random order.   Peak  heights  or  peak  areas are obtained for each
analyte.  Experience indicates that a linear calibration curve with zero
intercept is appropriate for each analyte.   Therefore, a response factor
for  each  analyte can be taken as  the slope of  the  best-fit regression
line.

      7.3.3 Daily Calibration. Analyze midpoint calibration standards, at
a minimum, at the beginning of the day, singly at  the midpoint of the run,
and singly after the  last sample of the day  (assuming a sample group of 10
samples or less).  Obtain  the response  factor for each  analyte from the
mean peak heights  or peak  areas  and compare  it with the response factor
obtained for the  initial calibration.  The mean  response factor for the
                             8330 - 8                         Revision 0
                                                          September 1994

-------
      daily calibration must  agree  within  ±15% of the response  factor  of the
      initial   calibration.    The same  criteria  is   required  for  subsequent
      standard  responses  compared  to  the  mean  response  of  the  triplicate
      standards beginning the day.   If this criterion  is not met, a new initial
      calibration must be obtained.

      7.4   HPLC Analysis

            7.4.1 Analyze the samples using the chromatographic conditions given
      in Sec.  7.2.  All positive measurements observed on the C-18 column must
      be confirmed by injection onto the CN column.

            7.4.2 Follow  Sec. 7.0  in  Method 8000  for  instructions  on  the
      analysis  sequence,  appropriate  dilutions,  establishing daily  retention
      time windows, and identification criteria.   Include a mid-level  standard
      after each  group of  10 samples  in  the analysis  sequence.   If  column
      temperature control  is not  employed,  special care must be taken to ensure
      that temperature shifts do not cause peak misidentification.

            7.4.3 Table 2 summarizes the estimated retention times on both C-18
      and CN columns for a number of analytes  analyzable using this method.  An
      example of the separation achieved by Column 1  is shown in Figure 1.

            7.4.4 Record the resulting peak sizes in peak heights  or area units.
      The use of peak heights is recommended to improve reproducibility of low
      level samples.

            7.4.5 Calculation of concentration is  covered  in Sec. 7.0 of Method
      8000.
8.0   QUALITY CONTROL

      8.1   Refer  to Chapter  One  for  specific  quality control  procedures.
Quality control to validate sample extraction is covered in Method 3500.

      8.2   Quality control required to validate  the  HPLC  system operation is
found in Method 8000, Sec. 8.0.

      8.3   Prior to preparation of stock solutions, acetonitrile, methanol, and
water blanks  should be run to determine  possible interferences  with  analyte
peaks.   If  the acetonitrile,  methanol,  or water  blanks  show  contamination,  a
different batch should be used.
9.0   METHOD PERFORMANCE

      9.1   Table 3 presents the  single laboratory precision based on data from
the analysis of  blind duplicates of four  spiked  soil  samples  and  four field
contaminated samples analyzed by seven laboratories.
                                   8330 - 9                         Revision 0
                                                                September 1994

-------
      9.2   Table 4 presents the multilaboratory  error  based  on data from the
analysis  of blind  duplicates  of  four  spiked soil  samples  and  four  field
contaminated samples analyzed by seven laboratories.

      9.3   Table  5  presents  the   multilaboratory   variance  of  the  high
concentration method for water based on data from nine laboratories.

      9.4   Table 6 presents multi laboratory recovery  data from the  analysis of
spiked soil samples by seven laboratories.

      9.5   Table 7 presents a comparison of method accuracy for soil  and aqueous
samples (high concentration method).

      9.6   Table 8 contains  precision  and accuracy  data  for the  salting-out
extraction method.
10.0  REFERENCES

1.    Bauer, C.F., T.F.  Jenkins,  S.M.  Koza,  P.M.  Schumacher,  P.H. Miyares and
      M.E.  Walsh  (1989).    Development  of  an  analytical  method  for  the
      determination of explosive residues in  soil.   Part 3.  Collaborative test
      results and final performance evaluation.  USA Cold Regions Research and
      Engineering Laboratory, CRREL Report 89-9.

2.    Grant,  C.L.,  A.D.  Hewitt  and  T.F. Jenkins   (1989)  Comparison  of  low
      concentration measurement capability estimates in trace  analysis:  Method
      Detection  Limits  and  Certified  Reporting  Limits.    USA  Cold  Regions
      Research and Engineering Laboratory, Special Report 89-20.

3.    Jenkins,   T.F.,   C.F.   Bauer,   D.C.   Leggett   and   C.L.   Grant  (1984)
      Reversed-phased HPLC method for analysis of  TNT, RDX,  HMX  and 2,4-DNT in
      munitions  wastewater.    USA  Cold  Regions  Research  and  Engineering
      Laboratory, CRREL Report 84-29.

4.    Jenkins, T.F. and  M.E.  Walsh  (1987) Development of an analytical method
      for explosive residues in  soil.  USA Cold Regions Research  and Engineering
      Laboratory, CRREL Report 87-7.

5.    Jenkins, T.F.,  P.H.  Miyares and ME. Walsh  (1988a)   An improved RP-HPLC
      method for determining nitroaromatics  and nitramines  in water.  USA Cold
      Regions Research and Engineering Laboratory, Special  Report 88-23.

6.    Jenkins,  T.F.   and  P.H.  Miyares  (1992)   Comparison of  Cartridge  and
      Membrane Solid-Phase  Extraction  with Salting-out  Solvent Extraction for
      Preconcentration  of Nitroaromatic   and   Nitramine  Explosives from Water.
      USA Cold Regions Research and  Engineering Laboratory,  Draft  CRREL Special
      Report.

7.    Jenkins,  T.F.,   P.W.   Schumacher,   M.E.  Walsh  and  C.F.   Bauer   (1988b)
      Development of  an analytical  method for  the  determination of explosive
      residues in soil.   Part II:  Further development and  ruggedness testing.
      USA Cold Regions Research and Engineering Laboratory, CRREL Report 88-8.

                                   8330 - 10                         Revision 0
                                                                September 1994

-------
8.    Leggett, D.C., T.F. Jenkins  and  P.M.  Miyares  (1990)  Salting-out solvent
      extraction  for  preconcentration  of neutral  polar organic  solutes  from
      water.  Analytical Chemistry, 62:  1355-1356.

9.    Miyares, P.H. and T.F. Jenkins (1990)  Salting-out solvent extraction for
      determining low  levels  of nitroaromatics and nitramines  in  water.   USA
      Cold Regions Research and Engineering  Laboratory, Special Report 90-30.


11.0  SAFETY

      11.1  Standard precautionary measures used  for handling  other organic
compounds should be sufficient  for  the safe handling  of the analytes targeted by
Method 8330.  The only extra caution that should be taken is when handling the
analytical standard neat material for the  explosives themselves and in rare cases
where soil or waste  samples are highly contaminated with the explosives.  Follow
the note for drying the neat materials at ambient temperatures.

      11.2  It is advisable  to  screen soil or waste samples using Method 8515 to
determine whether high concentrations of explosives are present.  Soil samples
as high  as  2% 2,4,6-TNT have  been safely ground.  Samples  containing higher
concentrations should not be ground in the mortar and pestle.   Method 8515 is for
2,4,6-TNT,  however, the other  nitroaromatics  will  also cause a  color  to  be
developed and provide  a rough  estimation  of  their concentrations.  2,4,6-TNT is
the analyte most often  detected in  high concentrations  in soil  samples.  Visual
observation of a soil  sample is also important when the sample  is taken from a
site expected to  contain  explosives.   Lumps of material that  have a chemical
appearance should be suspect and not ground.   Explosives  are generally a very
finely ground grayish-white material.
                                  8330  -  11                         Revision 0
                                                                September 1994

-------
           TABLE 1
ESTIMATED QUANTITATION LIMITS
Compounds
HMX
RDX
1,3,5-TNB
1,3-DNB
Tetryl
NB
2,4,6-TNT
4-Am-DNT
2-Am-DNT
2,6-DNT
2,4-DNT
2-NT
4-NT
3-NT
Water
Low-Level
-
0.84
0.26
0.11
-
-
0.11
0.060
0.035
0.31
0.020
-
-
-
(UQ/U
High-Level
13.0
14.0
7.3
4.0
4.0
6.4
6.9
-
-
9.4
5.7
12.0
8.5
7.9
Soil (mg/kg)
2.2
1.0
0.25
0.25
0.65
0.26
0.25
-
-
0.26
0.25
0.25
0.25
0.25
          8330  -  12
    Revision 0
September 1994

-------
                                    TABLE  2
        RETENTION  TIMES  AND CAPACITY FACTORS  ON  LC-18 AND LC-CN COLUMNS
Retention time
(min)
Compound
HMX
RDX
1,3,5-TNB
1,3-DNB
Tetry]
NB
2,4,6-TNT
4-Am-DNT
2-Am-DNT
2,6-DNT
2,4-DNT
2-NT
4-NT
3-NT
LC-18
2.44
3.73
5.11
6.16
6.93
7.23
8.42
8.88
9.12
9.82
10.05
12.26
13.26
14.23
LC-CN
8.35
6.15
4.05
4.18
7.36
3.81
5.00
5.10
5.65
4.61
4.87
4.37
4.41
4.45
Capacity
(k)
LC-18
0.49
1.27
2.12
2.76
3.23
3.41
4.13
4.41
4.56
4.99
5.13
6.48
7.09
7.68
factor
*
LC-CN
2.52
1.59
0.71
0.76
2.11
0.61
1.11
1.15
1.38
0.95
1.05
0.84
0.86
0.88
* Capacity factors are based  on  an  unretained  peak for nitrate at 1.71 min on
LC-18 and at 2.00 min on LC-CN.
                                   8330  -  13                         Revision 0
                                                                September 1994

-------
                                   TABLE 3
            SINGLE LABORATORY PRECISION OF METHOD FOR SOIL SAMPLES
                   Spiked Soils
RDX


1,3,5-TNB


1,3-DNB

Tetryl

2,4,6-TNT


2,4-DNT
                  Mean Cone.
             (mg/kg)        SD
                   %RSD
                    Field-Contaminated Soils
                             Mean Cone.
                   (mg/kg)     SD         %RSD
HMX
46
1.7
3.7
14
153
1.8
21.6
12.8
14.1
60


 8.6
46

 3.5

17

40


 5.0
1.4


0.4
1.9

0.14

3.1

1.4


0.17
 2.3


 4.6
 4.1

 4.0

17.9

 3.5


 3.4
104
877

  2.8
 72

  1.1

  2.3

  7.0
669

  1.0
12
29.6

 0.2
 6.0

 0.11

 0.41

 0.61
55

 0.44
11.5
 3.4

 7.1
 8.3

 9.8

18.0

 9.0
 8.2

42.3
                                   8330 - 14
                                                    Revision 0
                                                September 1994

-------
                       TABLE 4
  MULTILABORATORY ERROR OF METHOD FOR SOIL SAMPLES
Spiked Soils

HMX
RDX
1,3,5-TNB
1,3-DNB
Tetryl
2,4,6-TNT
2,4-DNT
Mean
(nig/ kg)
46
60
8.6
46
3.5
17
40
5.0
Cone.
SO
2.6
2.6
0.61
2.97
0.24
5.22
1.88
0.22
%RSD
5.7
4.4
7.1
6.5
6.9
30.7
4.7
4.4
Field-Contaminated Soils
(mg/kg)
14
153
104
877
2.8
72
1.1
2.3
7.0
669
1.0
Mean Cone.
SD °/.RSD
3.7
37.3
17.4
67.3
0.23
8.8
0.16
0.49
1.27
63.4
0.74
26.0
24.0
17.0
7.7
8.2
12.2
14.5
21.3
18.0
9.5
74.0
                       TABLE 5
MULTI LABORATORY VARIANCE OF METHOD FOR WATER SAMPLES8
Compounds
HMX
RDX
2,4-DNT
2,4,6-TNT
Mean Cone.
(MgA)
203
274
107
107
SD
14.8
20.8
7.7
11.1
%RSD
7.3
7.6
7.2
10.4
a Nine Laboratories
                      8330 -  15
    Revision 0
September 1994

-------
                                      TABLE 6
               MULTILABORATORY RECOVERY DATA FOR SPIKED SOIL SAMPLES
Laboratory
1
3
4
5
6
7
8
True Cone
Mean
Std Dev
% RSD
% Diff*
Mean %
Recovery
HMX
44.97
50.25
42.40
46.50
56.20
41.50
52.70
50.35
47.79
5.46
11.42
5.08
95
Concentration (/ig/g)
1,3,5- 1,3-
RDX TNB DNB
48.78
48.50
44.00
48.40
55.00
41.50
52.20
50.20
48.34
4.57
9.45
3.71
96
48.99
45.85
43.40
46.90
41.60
38.00
48.00
50.15
44.68
3.91
8.75
10.91
89
49.94
45.96
49.50
48.80
46.30
44.50
48.30
50.05
47.67
2.09
4.39
4.76
95
Tetryl
32.48
47.91
31.60
32.10
13.20
2.60
44.80
50.35
29.24
16.24
55.53
41.93
58
2,4,6-
TNT
49.73
46.25
53.50
55.80
56.80
36.00
51.30
50.65
49.91
7.11
14.26
1.46
98
2,4-
DNT
51.05
48.37
50.90
49.60
45.70
43.50
49.10
50.05
48.32
2.78
5.76
3.46
96
* Between true value and mean determined value.
                                      8330  -  16
    Revision 0
September 1994

-------
Analyte
                                    TABLE 7
          COMPARISON OF METHOD ACCURACY  FOR SOIL AND AQUEOUS SAMPLES
                          (HIGH CONCENTRATION  METHOD)
                                                Recovery (%)
Soil Method*
Aqueous Method**
2,4-DNT

2,4,6-TNT

RDX

HMX
    96.0

    96.8

    96.8

    95.4
      98.6

      94.4

      99.6

      95.5
*  Taken from Bauer et al. (1989), Reference 1.
** Taken from Jenkins et al. (1984), Reference 3.
                                   8330 -  17
                                Revision 0
                            September  1994

-------
                                   TABLE 8
      PRECISION AND ACCURACY DATA FOR THE SALTING-OUT EXTRACTION METHOD
Analyte
HMX
RDX
1,3,5-TNB
1,3-DNB
Tetryl
2,4,6-TNT
2-Am-DNT
2,4-DNT
1,2-NT
1,4-NT
1,3-NT
No. of Samples1
20
20
20
20
20
20
20
20
20
20
20
Precision
(% RSD)
10.5
8.7
7.6
6.6
16.4
7.6
9.1
5.8
9.1
18.1
12.4
Ave. Recovery
(%)
106
106
119
102
93
105
102
101
102
96
97
Cone. Range
(M9A)
0-1.14
0-1.04
0-0.82
0-1.04
0-0.93
0-0.98
0-1.04
0-1.01
0-1.07
0-1.06
0-1.23
1Reagent water
                                   8330  -  18
    Revision 0
September 1994

-------
 EXPLOSIVES ON A
 C18 COLUMN
            x
         5
                              I

'(5

	 	 k 4
1
6
i
a

id
	 u 	
	 U 	
EXPLOSIVES ON A
CN COLUMN
                                        1
                                                A
                                                         12       I 4
                             FIGURE  1
        CHROMATOGRAMS FOR COLUMNS  DESCRIBED  IN  Sec.  4.1.2.
       COURTESY  OF  U.S.  ARMY  CORPS OF  ENGINEERS, OMAHA,  NE.
                            8330 - 19
    Revision 0
September 1994

-------
                 METHOD  8330
  NITROAROMATICS  AND NITRAMINES  BY HIGH
PERFORMANCE LIQUID  CHROMATOGRAPHY  (HPLC)
        Low
Sailing Out
71 1 1 1 Add 251 3 got salt
and 1 L ot water sample to a
1 L vol flask Mix the contents

i >
71 1 12Add164mLof
acetonitnte (ACN) and stir
toe 1 5 mirts
i
i
71113 Transfer ACN layer
to 100 mL vol flask. Add 10 mU
of fresh ACN to 1 L flask and
stir Transfer 2nd portion and
combine with 1st in 100 mL flask.
I
i
71 1 1 4 Add 84 mL of sal
water to the ACN extract and stir
Transfer ACN extract to 10 mL
grad cylinder
i

7 1 1 1 5 Add 1 rnL of ACN to
1 00 mL vol flask Stir and
transfer to the 10 mL grad.
cylinder Record volume
Dilute 1 1 with reagent water
i

71116 Filter if turbid
Transfer to a vial lor
RP-HPLC analysis
                                     7111 Sample Filtration
                                      Place 5 mL sample in
                                     scintiltaion vtal. AddSmL
                                      methanol: shake, filler
                                   through 0 5 um filter Discard
                                    first 3 mL Retain remainder
                                         for use
                   8330  -  20
     Revision  0
September 1994

-------
                           METHOD  8330
                           (continued)
7121 Sample riomogemzation
Air dry sample at room Temp
or below  Avoid exposure to
direct sunlight Grind sample
in an acetonitrite nnsed mortar
  7122 Sample Extraction
71221
Place 2 g soil subsample.
10 mLs acetonitnle in 15 mL
glass vial Cap. vortex swirl.
place in room Temp or below
ultrasonic bath for 18 hrs
71222
Let sdn settle Add 5 ml
supernatant to 5 ml calcium
chloride soln  in 20 mL vial
Shake, let stand 15 mms
71223
Filter supernatant through
0 5 urn filter. Discard initial
3 mL. retain remainder
for analysis
7 2 Set Chromatographic Conditions
     7 3 Calibration of HPLC
 732
 Run working stds in triplicate
 Plot ng vs peak area or ht
 Curve should be linear with
 zero intercept
  733
  Analyze midrange calibration
  std at beginning, middle.
  and end of sample analyses
  Redo Section 7 3 1 if peak
  areas or hts do not agree
  to w/m w- 20% of initial
  calibration values
                                                         7 4 Sample Analysis
   741
   Analyze samples Confirm
   measurment w/injection onto
   CN column
                                                     743
                                                     Refer to Table 2 for typical
                                                     analyte retention times
                            8330   -   21
                               Revision  0
                         September  1994

-------
                                  METHOD 8331

                          TETRAZENE BY  REVERSE PHASE
                 HIGH PERFORMANCE LIQUID CHROMATOGRAPHY  (HPLC)
1.0   SCOPE AND APPLICATION

      1.1   This method is intended for the  analysis of tetrazene, an explosive
residue,  in  soil  and  water.    This  method  is  limited to  use  by  analysts
experienced  in handling  and  analyzing  explosive materials.    The following
compounds can be determined by this method:
      Compound                                              CAS No8
      Tetrazene                                             31330-63-9
      a  Chemical  Abstracts Service Registry number

      1.2   Tetrazene degrades rapidly in water and methanol at room temperature.
Special care must be taken to refrigerate or cool all solutions throughout the
analytical process.

      1.3   Tetrazene, in its dry form, is extremely explosive. Caution must be
taken during preparation of standards.

      1.4   The estimated quantitation limit (EQL) of Method 8331 for determining
the  concentration  of   tetrazene   is  approximately  7   jug/L  in  water  and
approximately 1 mg/kg in soil.

      1.5   This method  is restricted to use by or under  the supervision of
analysts  experienced  in the  use  of  HPLC,  skilled  in the  interpretation of
chromatograms, and experienced  in  handling  explosive materials.   Each analyst
must demonstrate the ability to generate acceptable results with this method.


2.0   SUMMARY OF METHOD

      2.1   A 10 mL water sample is filtered,  eluted  on a C-18  column using ion
pairing reverse phase HPLC, and quantitated at 280 nm.

      2.2   2  g  of  soil  are  extracted  with  55:45  v/v   methanol-water  and
1-decanesulfonic acid  on  a platform shaker, filtered,  and eluted on  a C-18 column
using ion pairing reverse phase HPLC,  and quantitated at 280 nm.
                                   8331 - 1                         Revision 0
                                                                September 1994

-------
3.0   INTERFERENCES

      3.1   No interferences are  known.  Tetrazene elutes early, however, and if
a computing integrator is used for peak quantification,  the baseline setting may
have to  be  set to exclude baseline aberrations.  Baseline setting is particularly
important at low concentrations of analyte.


4.0   APPARATUS AND MATERIALS

      4.1   HPLC system

            4.1.1   HPLC -  Pump  capable of achieving 4000 psi.

            4.1.2   100 jiL  loop  injector.

            4.1.3   Variable  or  fixed wavelength detector  capable  of reading
      280 nm.

            4.1.4   C-18  reverse phase HPLC  column,   25  cm x 4.6  mm  (5  nm)
      (Supelco LC-18,  or equivalent).

            4.1.5   Digital  integrator - HP 3390A (or  equivalent)

            4.1.6   Strip chart  recorder.

      4.2   Other apparatus

            4.2.1   Platform  orbital  shaker.

            4.2.2   Analytical balance - + 0.0001 g.

            4.2.3   Desiccator.

      4.3   Materials

            4.3.1   Injection  syringe  - 500 /uL.

            4.3.2   Filters - 0.5 nm Millex-SR and 0.5 /itm Millex-HV, disposable,
      or equivalent..

            4.3.3   Pipets  -  volumetric, glass,  Class  A.

            4.3.4   Scintillation vials -  20 mL, glass.

            4.3.5   Syringes  - 10 mL.

            4.3.6   Volumetric flasks, Class A  -  100 mL, 200 mL.

            4.3.7   Erlenmeyer flasks  with ground glass  stoppers  -  125  mL.
                                   8331 - 2                         Revision 0
                                                                September 1994

-------
      4.4   Preparation

            4.4.1   Prepare all materials as described in Chapter 4 for volatile
      organics.


5.0   REAGENTS

      5.1   HPLC grade chemicals shall be used  in all tests.  It  is intended that
all reagents shall  conform to  the  specifications of the Committee on Analytical
Reagents  of the  American  Chemical  Society,  where  such specifications  are
available.  Other grades may be used, provided  it is first ascertained that the
reagent is of sufficiently  high purity  to  permit  its  use without lowering the
accuracy of the determination.

      5.2   General

            5.2.1   Methanol,  CH3OH - HPLC grade.

            5.2.2   Organic-free reagent water - All references to water in this
      method refer to organic-free reagent water,  as defined in Chapter One.

            5.2.3   1-Decanesulfonic acid, sodium salt, C10H21S03Na - HPLC grade.

            5.2.4   Acetic  acid (glacial), CH3COOH - reagent grade.

      5.3   Standard Solutions

            5.3.1   Tetrazene  - Standard Analytical Reference Material.

            5.3.2   Stock standard solution - Dry tetrazene to constant weight
      in a vacuum desiccator in the dark.   (Tetrazene is extremely explosive in
      the dry state.   Do not dry more reagent  than  is necessary to prepare stock
      solutions.)   Place  about 0.0010  g  (weighed  to  0.0001 g)  into  a  100-ml
      volumetric flask and dilute to volume with methanol.  Invert flask several
      times until  tetrazene is dissolved.   Store  in freezer at  -10°C.   Stock
      solution is about 100 mg/L.   Replace stock standard solution every week.

            5.3.3   Intermediate standard  solutions

                    5.3.3.1   Prepare a 4  mg/L standard  by  diluting  the stock
            solution 1/25 v/v  with methanol.

                    5.3.3.2   Pipet 0.5,  1.0, 2.0, 5.0, 10.0,  and 20.0 mL of the
            4 mg/L standard solution into 6 separate 100 mL volumetric flasks,
            and make up to volume with methanol.  Pipet  25.0 mL of  the  4 mg/L
            standard  solution  into  a 50 mL volumetric  flask,  and make  up  to
            volume with methanol.   This results in intermediate  standards  of
            about 0.02, 0.04,  0.08,  0.2, 0.4,  0.8,  2 and 4 mg/L.

                    5.3.3.3   Cool  immediately on preparation in refrigerator or
            ice bath.
                                   8331 - 3                         Revision 0
                                                                September 1994

-------
            5.3.4   Working standard solutions

                    5.3.4.1   Inject 4 mL of each of the intermediate standard
            solutions into 6.0 ml of water.  This results in concentrations of
            about 0.008, 0.016,  0.032,  0.08, 0.16, 0.3, 0.8 and 1.6 mg/L.

                    5.3.4.2   Cool immediately on preparation in refrigerator or
            ice bath.

      5.5   QC spike concentrate solution

            5.5.1   Dry tetrazene to constant weight in a vacuum desiccator in
      the dark.  (Tetrazene  is extremely explosive in the dry state.  Do not dry
      any more than necessary  to  prepare standards.)   Place about  0.0011  g
      (weighed to 0.0001 g)  into a 200-ml  volumetric flask and dilute to volume
      with methanol.  Invert flask several times until  tetrazene is dissolved.
      Store in  freezer  at -10°C.   QC  spike concentrate solution  is  about 55
      mg/L.  Replace stock standard solution every week.

            5.5.2   Prepare spiking solutions, at  concentrations appropriate to
      the concentration  range  of the samples being analyzed, by diluting the QC
      spike  concentrate  solution  with  methanol.   Cool  on  preparation  in
      refrigerator or ice bath.

      5.6   Eluent

            5.6.1   To  make  about  1  liter  of  eluent,   add   2.44   g  of
      1-decanesulfonic acid, sodium salt  to 400/600 v/v methanol/water, and add
      2.0 ml of glacial  acetic acid.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material to this Chapter,  Organic Analytes,
Section 4.1.

      6.2   Samples must be collected  and stored  in  glass  containers.   Follow
conventional sampling procedures.

      6.3   Samples must be kept below 4°C from the time of collection through
analysis.


7.0   PROCEDURE

      7.1   Sample Preparation

            7.1.1   Filtration  of Water Samples

                    7.1.1.1   Place a  10  mL portion  of each water sample in a
            syringe and  filter through  a  0.5 /urn Millex-HV filter unit.  Discard
            first 5 mL of filtrate, and retain 5 mL for analysis.
                                   8331 - 4                         Revision 0
                                                                September 1994

-------
      7.1.2    Extraction  and  Filtration of Soil Samples

               7.1.2.1   Determination of sample %  dry weight - In certain
      cases, sample results are desired based on dry-weight basis.  When
      such data  is  desired,  a portion of sample  for  this determination
      should be  weighed out  at  the same time  as the portion  used for
      analytical determination.

               WARNING:  The drying oven should be contained in a hood or
                        vented.  Significant  laboratory contamination may
                        result  from  a heavily  contaminated  hazardous
                        waste sample.

                  7.1.2.1.1   Immediately after weighing  the  sample for
               extraction,  weigh  5-10  g  of the  sample  into  a  tared
               crucible.   Determine  the  % dry  weight  of the  sample by
               drying overnight  at  105°C.  Allow to  cool  in a desiccator
               before weighing:

                  % dry weight = q of dry sample x 100
                                     g of sample

               7.1.2.2   Weigh 2 g soil subsamples into 125 ml Erlenmeyer
      flasks with ground glass stoppers.

               7.1.2.3   Add  50  ml  of 55/45   v/v  methanol-water  with
      1-decanesulfonic acid, sodium salt added to make a 0.1 M solution.

               7.1.2.4   Vortex for 15 seconds.

               7.1.2.5   Shake for 5 hr at 2000 rpm on platform shaker.

               7.1.2.6   Place a 10  ml portion of each soil sample extract
      in a  syringe  and filter through  a  0.5 /jm  Millex-SR  filter unit.
      Discard first 5 ml of filtrate, and retain 5 ml for analysis.

7.2   Sample Analysis

      7.2.1    Analyze  the samples  using  the chromatographic  conditions
given in Section 7.2.1.1.  Under these conditions, the retention time of
tetrazene is 2.8 min.   A sample  chromatogram,  including  other compounds
likely  to   be  present  in  samples  containing  tetrazene,  is  shown  in
Figure 1.

               7.2.1.1   Chromatographic Conditions

               Solvent:              0.01  M  1-decanesulfonic  acid,  in
                                    acidic methanol/water (Section 5.5)
               Flow  rate:            1.5 mL/min
               Injection  volume:     100 /xL
               UV Detector:          280 nm
                             8331 - 5                         Revision 0
                                                          September 1994

-------
      7.3   Calibration of HPLC

            7.3.1    Initial  Calibration  -    Analyze  the  working  standards
      (Section 5.3.4), starting with the  0.008 mg/L  standards and ending with
      the 0.30 mg/L standard.   If the percent relative standard deviation (%RSD)
      of the mean response  factor (RF) for each analyte  does not exceed 20%, the
      system is calibrated  and  the analysis of samples may proceed.  If the %RSD
      for any analyte  exceeds  20%,  recheck the system  and/or recalibrate with
      freshly prepared calibration solutions.

            7.3.2    Continuing Calibration - On a daily basis, inject 250 /j,l of
      stock standard  into  20 ml water.   Keep solution  in  refrigerator until
      analysis.  Analyze  in triplicate (by overfilling loop) at the beginning of
      the day, singly after each  five samples,  and singly after the last sample
      of the  day.   Compare response factors  from the  mean peak  area  or peak
      height  obtained over  the  day  with  the  response  factor  at  initial
      calibration.  If these values do  not agree within 10%,  recalibrate.


8.0   QUALITY CONTROL

      8.1   Refer to Chapter One for specific quality control procedures.

      8.2   Prior to preparation of stock solutions, methanol  should be analyzed
to determine  possible  interferences with  the  tetrazene peak.   If the methanol
shows contamination, a different batch  of methanol should be used.

      8.3   Method Blanks

            8.3.1    Method blanks  for  the analysis of  water  samples  should be
      organic-free reagent  water carried through all sample storage and handling
      procedures.

            8.3.2    Method blanks  for  the analysis of soil  samples  should be
      uncontaminated soil  carried through  all  sample  storage,  extraction, and
      handling procedures.


9.0   METHOD  PERFORMANCE

      9.1   Method 8331 was tested in a laboratory over a period of four days.
Spiked organic-free reagent water and standard soil  were analyzed in duplicate
each  day for  four days.   The  HPLC  was  calibrated  daily  according  to the
procedures given in Section 7.1. Method performance data are presented in Tables
1 and 2.


10.0  REFERENCES

1.    Walsh,  M.E.,  and  T.F.  Jenkins,   "Analytical   Method  for  Determining
      Tetrazene  in Water,"  U.S. Army Corps of Engineers,  Cold Regions Research
      & Engineering Laboratory, Special Report 87-25, 1987.
                                   8331 - 6                         Revision 0
                                                                September 1994

-------
2.    Walsh,  M.E.,  and  T.F.   Jenkins,   "Analytical  Method  for  Determining
      Tetrazene in Soil," U.S.  Army Corps  of  Engineers, Cold Regions Research &
      Engineering Laboratory, Special Report 88-15, 1988.


11.0  SAFETY

      11.1   Standard precautionary measures  used  for handling  other organic
compounds should  be  sufficient for safe handling of  the analytes  targeted by
Method 8331.
                                   8331  -  7
    Revision 0
September 1994

-------
               FIGURE  1
   16
   12
^m


8   8
                           TNT
                HMX
                           Tttroztnt
   O1-
                    0.064
               Absorbonct Units
              8331 - 8
    Revision 0
September  1994

-------
            TABLE 1.
METHOD PERFORMANCE,  WATER MATRIX
Spike
Cone.
(M9A)
0.00



7.25



14.5



29



72.5



145



290



725



OVERALL
Avq % Recovery
Replicate
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery

Day 1
0.0
NA
0.0
NA
8.9
122
6.6
91
14.6
101
14.8
102
31.8
110
29.5
102
71.1
98
71.2
98
140.6
97
138.5
96
289.4
100
282.0
97
737.6
102
700.2
97

Day 2
0.0
NA
0.0
NA
7.8
108
9.9
' 137
14.6
101
14.1
97
30.0
103
29.7
102
73.6
102
71.3
98
143.8
99
140.8
97
288.5
99
284.2
98
707.2
98
695.8
96

Day 3
0.0
NA
0.0
NA
7.4
102
8.5
117
13.8
95
14.1
98
30.8
106
30.4
105
75.7
104
70.7
98
144.7
100
140.9
97
291.0
100
281.9
97
714.3
99
714.2
99

Average
Day 4
0.0
NA
0.0
NA
9.4
130
6.7
92
14.6
101
15.2
105
28.7
99
30.7
106
73.9
102
71.6
99
142.1
98
136.9
94
289.8
100
282.5
97
722.0
100
716.3
99

% Recovery

NA

NA

116

109

99

100

105

104

101

98

98

96

100

97

99

97
102
            8331  -  9
    Revision 0
September 1994

-------
            TABLE 2
METHOD PERFORMANCE, SOIL MATRIX
Spike
Cone.
(M9/L)
0.00



1.28



2.56



5.12



12.8



25.6



OVERALL
Avq % Recovery
Replicate
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Repl icate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery
Replicate 1
% Recovery
Replicate 2
% Recovery

Day 1
0.0
NA
0.0
NA
0.6
49
1.2
92
1.4
56
1.5
59
2.9
57
3.0
58
7.8
61
8.0
62
17.2
67
16.7
65

Day 2
0.0
NA
0.0
NA
0.9
73
0.7
56
1.5
58
2.0
79
3.0
58
3.0
59
7.6
59
8.4-
66
16.7
65
16.8
66

Day 3
0.0
NA
0.0
NA
0.6
48
0.8
63
1.6
61
1.4
56
2.9
56
3.5
69
7.8
61
7.7
60
17.4
68
17.6
69

Average
Day 4 %
0.0
NA
0.0
NA
1.0
74
0.7
56
1.6
61
1.3
50
2.9
56
3.1
60
8.1
63
8.2
64
17.3
68
17.2
67

Recovery

NA

NA

61

67

59

61

57

61

61

63

67

67
62
           8331  -  10
    Revision 0
September 1994

-------
                                     METHOD  8331
                           TETRAZENE BY REVERSE PHASE
               HIGH  PERFORMANCE  LIQUID  CHROMATOGRAPHY  (HPLC)
        Start
   7.1 .1 Filter 10 ml
 water sample; discard
first 5 mL; analyze last 5.
  7.1.2.1  Determine
    % dry weight.
  7.1.2.2 - 7.1.2.5
   Extract 2 g soil
 with 50 mL solvent.
  7.1.2.6 Filter 10 mL
 extract; discard 5 mL;
  analyze last 5 mL.
 7.2 Analyze samples
using chromatographic
    conditions in
   Section 7.2.1.1.
                                     7.3.1  Initial calibration:
                                        Analyze working
                                           standards
                                        (Section 5.3.3).
                                                                       7.3.1 Recheck system/
                                                                        recalibrate with new
                                                                        calibration solution.
                                              7.3.2
                                            Continuing
                                            Calibration.
                                               Stop
                                      8331  -  11
                                           Revision  0
                                      September 1994

-------
4.3  DETERMINATION OF ORGANIC ANALYTES

     4.3.4  FOURIER TRANSFORM INFRARED METHODS

         The following method is included in this section:

         Method 8410:      Gas  Chromatography/Fourier  Transform  Infrared
                           (GC/FT-IR)    Spectrometry   for    Semi volatile
                           Organics:  Capillary Column
                                 FOUR -  13                        RevisionZ
                                                             September 1994

-------
                                  METHOD  8410

                 GAS CHROMATOGRAPHY/FOURIER TRANSFORM  INFRARED
              (GC/FT-IR) SPECTROMETRY FOR SEMIVOLATILE ORGANICS:
                               CAPILLARY COLUMN
1.0   SCOPE AND APPLICATION

      1.1   This method covers the automated identification, or compound class
assignment  of  unidentifiable compounds,  of solvent  extractable  semivolatile
organic  compounds which  are  amenable  to  gas  chromatography,   by  GC/FT-IR.
GC/FT-IR can be  a useful  complement  to  GC/MS  analysis (Method 8270).   It is
particularly well  suited for  the identification of specific isomers that are not
differentiated using GC/MS.   Compound  class  assignments are made using infrared
group  absorption   frequencies.    The  presence  of  an  infrared   band  in  the
appropriate group  frequency  region may  be  taken as evidence  of  the possible
presence of a particular compound class, while its absence may be construed as
evidence that the  compound class in question is not present.  This  evidence will
be further strengthened by the presence of confirmatory group frequency bands.
Identification limits of the  following compounds  have been demonstrated by this
method.
      Compound Name
CAS No.'
Acenaphthene
Acenaphthylene
Anthracene
Benzo( a) anthracene
Benzo(a)pyrene
Benzoic acid
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl } ether
Bis(2-ethylhexyl) phthalate
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
4-Chloroanil ine
4 -Chloro- 3 -methyl phenol
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenol
4-Chlorophenyl phenyl ether
Chrysene
Dibenzofuran
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1,4-Dichlorobenzene
2,4-Dichlorophenol
83-32-9
208-96-8
120-12-7
56-55-3
50-32-8
65-85-0
111-91-1
111-44-4
39638-32-9
117-81-7
101-55-3
85-68-7
106-47-8
59-50-7
91-58-7
95-57-8
106-48-9
7005-72-3
218-01-9
132-64-9
84-74-2
95-50-1
541-73-1
106-46-7
120-83-2
                                   8410 - 1
                  Revision 0
              September 1994

-------
      Compound Name                               CAS No.a


      Dimethyl phthalate                          131-11-3
      Diethyl  phthalate                            84-66-2
      4,6-Dinitro-2-methylphenol                   534-52-1
      2,4-Dinitrophenol                            51-28-5
      2,4-Dinitrotoluene                          121-14-2
      2,6-Dinitrotoluene                          606-20-2
      Di-n-octyl  phthalate                        117-84-0
      Di-n-propyl  phthalate                       131-16-8
      Fluoranthene                                206-44-0
      Fluorene                                     86-73-7
      Hexachloroberizene                           118-74-1
      1,3-Hexachlorobutadiene                      87-68-3
      Hexachlorocyclopentadiene                    77-47-4
      Hexachloroethane                             67-72-1
      Isophorone                                   78-59-1
      2-Methylnaphthalene                          91-57-6
      2-Methylphenol                                95-48-7
      4-Methylphenol                               106-44-5
      Naphthalene                                   91-20-3
      2-Nitroaniline                               88-74-4
      3-Nitroaniline                               99-09-2
      4-Nitroaniline                              100-01-6
      Nitrobenzene                                 98-95-3
      2-Nitrophenol                                88-75-5
      4-Nitrophenol                               100-02-7
      N-Nitrosodimethylamine                       62-75-9
      N-Nitrosodiphenylamine                       86-30-9
      N-Nitroso-di-n-propylamine                  621-64-7
      Pentachlorophenol                            87-86-5
      Phenanthrene                                 85-01-8
      Phenol                                      108-95-2
      Pyrene                                      129-00-0
      1,2,4-Trichlorobenzene                      120-82-1
      2,4,5-Trichlorophenol                        95-95-4
      2,4,6-Trichlorophenol                        88-06-2


      a  Chemical  Abstract  Services  Registry Number.

      1.2   This method is  applicable  to the determination of most extractable,
semivolatile-organic compounds  in wastewater,  soils and sediments,  and solid
wastes.  Benzidine can be subject to losses  during solvent concentration and GC
analysis;  a-BHC,   /3-BHC,  Endosulfan  I  and  II,  and  Endrin  are  subject  to
decomposition under the  alkaline conditions  of  the  extraction  step;  Endrin is
subject to decomposition during GC analysis; and Hexachlorocyclopentadiene and
N-Nitrosodiphenylamine may decompose during extraction and  GC analysis.  Other
extraction and/or  instrumentation procedures should be considered for unstable
analytes.


                                   8410 - 2                         Revision 0
                                                                September 1994

-------
      1.3   The identification limit of this method may depend strongly upon the
level and  type of gas  chromatographable  (GC)  semivolatile extractants.   The
values listed in Tables  1 and 2 represent the minimum quantities of semivolatile
organic compounds which have been identified by the specified GC/FT-IR system,
using this  method and under routine environmental analysis conditions.  Capillary
GC/FT-IR wastewater identification limits of 25 M9/L  may be achieved for weak
infrared absorbers  with this  method,  while the  corresponding  identification
limits for  strong infrared  absorbers is 2 M9/L.   Identification limits for other
sample matrices can  be calculated from the wastewater  values  after choice of the
proper sample workup procedure (see Sec. 7.1).


2.0   SUMMARY OF METHOD

      2.1   Prior to  using this method,  the  samples  should be prepared  for
chromatography  using the  appropriate  sample preparation  and  cleanup methods.
This  method describes  chromatographic  conditions that  will   allow for  the
separation   of  the  compounds in the extract and uses  FT-IR  for  detection  and
quantitation of the target analytes.


3.0   INTERFERENCES

      3.1   Glassware and  other sample  processing  hardware  must  be thoroughly
cleaned to  prevent contamination  and misinterpretation.  All of these materials
must be demonstrated to be free from interferences under the conditions of the
analysis  by  running  method   blanks.    Specific  selection  of  reagents  or
purification of solvents by distillation in all-glass systems may be required.

      3.2   Matrix interference will  vary considerably from source to source,
depending upon the diversity of the residual  waste  being sampled.  While general
cleanup  techniques  are  provided as  part of this  method,  unique  samples  may
require additional cleanup  to isolate the analytes of interest from interferences
in order to achieve maximum sensitivity.

      3.3   4-Chlorophenol  and  2-nitrophenol are subject to interference from co-
eluting compounds.

      3.4   Clean all glassware as soon as possible  after use by rinsing with the
last solvent used.   Glassware should  be sealed/stored  in  a clean  environment
immediately  after  drying  to prevent   any accumulation   of  dust  or  other
contaminants.
4.0   APPARATUS AND MATERIALS

      4.1   Gas  Chromatographic/Fourier   Transform   Infrared   Spectrometric
Equipment

            4.1.1  Fourier Transform-Infrared  Spectrometer  -  A  spectrometer
      capable of collecting at least one scan set per second at 8 cm"1 resolution
      is  required.    In  general,   a  spectrometer  purchased  after  1985,  or
      retrofitted to meet  post-1985 FT-IR improvements, will  be  necessary to


                                   8410 - 3                         Revision 0
                                                                September 1994

-------
            meet  the detection limits of this protocol.  A state-of-the-art A/D
            converter  is  required,  since  it  has  been  shown  that  the  signal-to-
            noise ratio  of  single beam  GC/FT-IR  systems is  A/D  converter
            1imited.

            4.1.2 GC/FT-IR Interface - The interface should  be lightpipe volume-
      optimized for the selected chromatographic  conditions  (lightpipe volume of
      100-200 fj,L  for capillary columns).   The shortest possible  inert transfer
      line (preferably fused  silica) should be used to interface  the end of the
      chromatographic  column  to the  lightpipe.   If  fused  silica  capillary
      columns are employed, the  end of the GC column  can serve as  the transfer
      line  if  it is  adequately  heated.   It has  been  demonstrated that  the
      optimum lightpipe volume is equal to the full width at half height of the
      GC  eluate peak.

            4.1.3 Capillary  Column  -  A  fused   silica  DB-5  30  m  x 0.32  mm
      capillary column with 1.0  jum film  thickness (or equivalent).

            4.1.4 Data Acquisition - A computer system dedicated to the GC/FT-IR
      system to  allow the continuous  acquisition  of  scan  sets  for  a  full
      chromatographic  run.  Peripheral data storage systems should be available
      (magnetic  tape  and/or  disk) for  the  storage  of  all  acquired  data.
      Software should  be available to allow the acquisition  and storage of every
      scan set to locate  the  file numbers and transform high S/N  scan sets, and
      to  provide  a real time  reconstructed chromatogram.

            4.1.5 Detector - A cryoscopic,  medium-band  HgCdTe  (MCT) detector
      with the  smallest practical focal area.   Typical narrow-band MCT detectors
      operate  from  3800-800  cm"1,  but  medium-band  MCT detectors   can  reach
      650 cm"1.   A 750 cm"1 cutoff  (or lower)  is  desirable  since  it  allows the
      detection of typical  carbon-chlorine  stretch and  aromatic  out-of-plane
      carbon-hydrogen  vibrations of environmentally  important organo-chlorine
      and polynuclear aromatic  compounds.   The  MCT  detector  sensitivity (D)*
      should be > 1  x  1010  cm.

            4.1.6  Lightpipe  - Constructed of  inert materials,  gold coated, and
      volume-optimized for  the  desired   chromatographic  conditions  (see  Sec.
      7.3).

            4.1.7 Gas Chromatograph  -   The  FT-IR   spectrometer  should  be
      interfaced  to  a  temperature programmable gas chromatograph equipped with
      a Grob-type (or  equivalent) purged  splitless injection system suitable for
      capillary glass  columns or an on-column injector system.

            A short, inert  transfer line should  interface the gas chromatograph
      to  the  FT-IR  lightpipe and,  if applicable,  to  the GC detector.   Fused
      silica GC columns may  be directly  interfaced to  the  lightpipe inlet and
      outlet.

      4.2   Dry Purge  Gas  -  If the  spectrometer is the  purge-type,  provisions
should be made to provide a suitable continuous  source of dry  purge-gas to the
FT-IR spectrometer.
                                   8410 - 4                         Revision 0
                                                                September 1994

-------
      4.3   Dry Carrier  Gas  -  The carrier  gas  should  be  passed  through  an
efficient cartridge-type drier.

      4.4   Syringes - 1-jj.l,  10-juL.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be used in  all tests. Unless
otherwise  indicated,  it is  intended  that all reagents  shall conform  to the
specifications of the Committee on Analytical Reagents of the American Chemical
Society, where  such specifications are  available.  Other grades may  be used,
provided it is first ascertained that the reagent  is  of sufficiently  high purity
to permit its use without lessening the accuracy of the determination.

      5.2   Organic-free reagent water.   All  references to water in  this method
refer to organic-free reagent water,  as defined in Chapter One.

      5.3   Solvents

            5.3.1  Acetone,  CH3COCH3 - Pesticide quality, or equivalent.

            5.3.2  Methylene chloride, CH2C12  - Pesticide quality, or  equivalent.

      5.4   Stock Standard  Solutions  (1000  mg/L)  - Standard  solutions  can  be
prepared from pure standard materials or purchased as a certified solution.

            5.4.1  Prepare stock standard solutions by accurately weighing 0.1000
      + 0.0010 g of pure material.  Dissolve the material in pesticide quality
      acetone  or  other  suitable   solvent and  dilute to  volume  in a  100  ml
      volumetric flask.  Larger volumes  can  be used at  the convenience of the
      analyst.  When compound purity is assayed to be 96 percent or greater, the
      weight may be  used without  correction  to  calculate the  concentration  of
      the stock standard. Commercially  prepared stock standards may be used at
      any  concentration  if   they  are  certified  by  the  manufacturer or  by  an
      independent source.

            5.4.2  Transfer  the stock standard solutions into bottles  with Teflon
      lined screw-caps.  Store  at 4°C and protect  from  light.   Stock standard
      solutions  should  be  checked  frequently  for  signs  of degradation  or
      evaporation,  especially just prior to preparing calibration standards from
      them.

            5.4.3  Stock standard solutions  must  be replaced after  6 months  or
      sooner if comparison  with quality  control  reference samples  indicates a
      problem.

      5.5   Calibration Standards  and Internal Standards - For  use in situations
where GC/FT-IR will  be used for  primary quantitation of  analytes  rather than
confirmation of GC/MS identification.

            5.5.1  Prepare calibration  standards  that contain  the compounds  of
      interest, either singly  or mixed  together.   The  standards  should  be


                                   8410 - 5                          Revision 0
                                                                September 1994

-------
      prepared at concentrations that will completely bracket the working range
      of  the chromatographic  system  (at  least  one  order of  magnitude  is
      suggested).

            5.5.2  Prepare internal  standard solutions.    Suggested  internal
      standards  are  1-Fluoronaphthalene,  Terphenyl,  2-Chlorophenol,  Phenol,
      Bis(2-chloroethoxy)methane, 2,4-Dichlorophenol, Phenanthrene, Anthracene,
      and Butyl  benzyl phthalate.  Determine the  internal standard concentration
      levels from the minimum identifiable quantities.   See Tables 1 and 2.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the introductory material to  this chapter, Organic Analytes, Sec.
4.1.


7.0   PROCEDURE

      7.1   Sample Preparation  - Samples must be  prepared by one  of the following
methods prior to GC/FT-IR analysis.

      Matrix                        Methods

      Water                         3510, 3520
      Soil/sediment                 3540, 3541,  3550
      Waste                         3540, 3541,  3550, 3580

      7.2   Extracts  may be cleaned up by Method 3640,  Gel-Permeation Cleanup.

      7.3   Initial  Calibration - Recommended GC/FT-IR  conditions:

      Scan time:                                At least 2 scan/sec.
      Initial column temperature and hold time:  40°C  for 1 minute.
      Column temperature program:               40-280°C at 10°C/min.
      Final column temperature hold:            280°C.
      Injector temperature:                     280-300°C.
      Transfer line temperature:                270°C.
      Lightpipe:                                280°C.
      Injector:                                 Grob-type,  splitless  or  on-
                                                column.
      Sample volume:                             2-3 yl.
      Carrier gas:                              Dry helium at about 1 mL/min.

      7.4   With an oscilloscope, check the detector centerburst  intensity versus
the manufacturer's specifications.  Increase the source voltage, if necessary,
to  meet these  specifications.    For  reference  purposes,  laboratories should
prepare a plot of time  versus detector voltage over at least a 5 day period.

      7.5   Capillary Column  Interface  Sensitivity  Test  -  Install  a 30  m  x
0.32 mm  fused  silica  capillary  column  coated   with  1.0  jum  of  DB-5   (or
equivalent).   Set  the lightpipe and transfer lines  at  280°C,  the injector at
225°C and the GC detector at  280°C  (if  used).  Under splitless Grob-type or on-


                                   8410  - 6                         Revision  0
                                                                September  1994

-------
column injection conditions,  inject  25  ng of nitrobenzene, dissolved in 1 jul_ of
methylene chloride.  The nitrobenzene should be identified by the on-line library
software search within  the first five hits (nitrobenzene  should  be contained
within the search library).

      7.6   Interferometer -  If  the interferometer is  air-driven,  adjust the
interferometer drive air pressure to manufacturer's specifications.

      7.7   MCT Detector Check -  If the centerburst intensity is 75 percent or
less of the mean intensity  of the plot maximum  obtained by the procedure of Sec.
7.4, install a new  source  and check the MCT  centerburst with  an  oscilloscope
versus the manufacturer's  specifications (if  available).   Allow at least five
hours of new source operation before data acquisition.

      7.8   Frequency Calibration -  At the  present time,  no  consensus  exists
within the spectroscopic community on a  suitable frequency reference  standard for
vapor-phase FT-IR.  One reviewer has  suggested  the use of indene as an on-the-fly
standard.

      7.9   Minimum  Identifiable  Quantities  -  Using  the   GC/FT-IR operating
parameters specified in  Sec.  7.3, determine the minimum identifiable quantities
for the compounds of interest.

            7.9.1   Prepare  a plot of  lightpipe temperature versus MCT centerburst
      intensity (in volts  or  other  vertical  height units).  This  plot  should
      span the  temperature  range between ambient and the  lightpipe thermal limit
      in increments  of about 20°C.  Use  this plot for daily QA/QC  (see Sec. 8.4).
      Note that modern GC/FT-IR interfaces (1985 and later)  may have eliminated
      most of this temperature effect.

      7.10  GC/FT-IR Extract  Analysis

            7.10.1      Analysis  - Analyze the dried methylene chloride extract
      using the chromatographic conditions  specified in Sec. 7.3 for capillary
      column interfaces.

            7.10.2      GC/FT-IR  Identification  -  Visually  compare the analyte
      infrared (IR)  spectrum  versus the search  library spectrum  of  the most
      promising on-line  library  search  hits.    Report,  as  identified,  those
      analytes  with  IR frequencies for the five (maximum number) most intense IR
      bands (S/N > 5) which are within  +  5.0 cm"1 of the corresponding bands in
      the library  spectrum.  Choose  IR  bands which are sharp and well resolved.
      The software used  to  locate spectral peaks should employ the peak "center
      of gravity"  technique.   In addition, the  IR frequencies of the analyte and
      library spectra should  be determined with  the same computer  software.

            7.10.3      Retention Time  Confirmation - After  visual comparison of
      the analyte  and library  spectrum  as described in Sec.  7.10.2, compare the
      relative  retention times  (RRT) of the analyte and  an authentic standard of
      the most  promising  library search hit.   The  standard and  analyte RRT
      should agree within + 0.01 RRT units when both are determined at the same
      chromatographic conditions.
                                   8410 - 7                         Revision 0
                                                                September 1994

-------
            7.10.4       Compound  Class  or  Functionality Assignment  -  If  the
      analyte cannot be unequivocally identified,  report its  compound class or
      functionality.  See Table 3 for gas-phase group frequencies to be used as
      an aid for compound class assignment.  It should be noted that FT-IR gas-
      phase group stretching  frequencies are 0-30  cm"1 higher  in frequency than
      those of the  condensed  phase.

            7.10.5       Quantitation  - This protocol can be used to confirm GC/MS
      identifications, with subsequent quantitation.   Two FT-IR quantitation and
      a supplemental GC detector  technique are also  provided.

                   7.10.5.1    Integrated Absorbance  Technique  - After analyte
            identification,   construct   a  standard   calibration    curve   of
            concentration  versus  integrated   infrared  absorbance.     For  this
            purpose, choose for integration only those FT-IR scans which are at
            or above the peak half-height.  The calibration curve should span at
            least one order of magnitude and  the  working  range should bracket
            the analyte concentration.

                   7.10.5.2    Maximum Absorbance  Infrared  Band  Technique  -
            Following analyte identification,  construct a standard  calibration
            curve of concentration versus maximum  infrared band intensity.  For
            this purpose,  choose  an  intense,   symmetrical  and  well  resolved IR
            absorbance band.

                   (Note that IR transmission  is not proportional to concentra-
            tion).   Select  the FT-IR scan with the  highest absorbance to plot
            against concentration.   The  calibration  curve should span at least
            one order  of magnitude  and  the working  range should  bracket  the
            analyte  concentration.      This   method   is  most  practical   for
            repetitive, target compound  analyses.   It is  more sensitive than the
            integrated absorbance  technique.

                   7.10.5.3    Supplemental  GC  Detector Technique  -  If a  GC
            detector is used in tandem  with the FT-IR detector,  the following
            technique may  be  used:  following  analyte identification, construct
            a standard calibration curve of concentration versus integrated peak
            area.   The calibration  curve  should  span  at  least one  order of
            magnitude  and   the   working  range  should  bracket  the  analyte
            concentration.  This method  is most practical for repetitive, target
            compound analyses.


8.0   QUALITY CONTROL

      8.1   Refer  to Chapter One  for  specific   quality  control  procedures.
Quality control to validate sample extraction  is  covered in Method  3500 and in
the extraction method utilized.  If extract cleanup was performed, follow the QC
in Method 3600 and  in the specific cleanup method.

      8.2   One Hundred Percent Line Test - Set the  GC/FT-IR operating conditions
to those employed for the Sensitivity Test (see Sec. 7.5).  Collect 16  scans over
the entire detector spectral  range.   Plot  the  test and measure the peak-to-peak


                                   8410 - 8                         Revision 0
                                                                September 1994

-------
noise between 1800 and  2000 cm"1.  This noise should be < 0.15%.  Store this plot
for future reference.

      8.3   Single Beam Test  - With the GC/FT-IR at analysis conditions, collect
16 scans  in  the  single beam  mode.   Plot the co-added  file  and  compare with a
subsequent file acquired in the same fashion several  minutes  later.  Note if the
spectrometer  is  at  purge equilibrium.    Also  check  the  plot  for signs  of
deterioration of the lightpipe potassium bromide windows.  Store this plot for
future reference.

      8.4   Align  Test  -  With  the   lightpipe  and  MCT detector  at  thermal
equilibrium, check the  intensity of the centerburst versus the signal temperature
calibration curve.  Signal  intensity deviation from the predicted intensity may
mean thermal equilibrium has not yet  been  achieved,  loss of detector coolant,
decrease  in  source  output,  or  a  loss  in signal  throughput resulting  from
lightpipe deterioration.

      8.5   Mirror Alignment  -  Adjust  the interferometer mirrors to attain the
most  intense signal.    Data collection  should not  be  initiated until  the
interferogram is stable.   If necessary, align the mirrors  prior to each GC/FT-IR
run.

      8.6   Lightpipe - The lightpipe  and lightpipe windows should be protected
from moisture and  other  corrosive  substances at  all  times.   For this purpose,
maintain the lightpipe  temperature  above the maximum GC program temperature but
below its thermal  degradation limit.  When  not  in  use,  maintain  the lightpipe
temperature slightly above ambient.  At all  times, maintain a  flow of dry, inert,
carrier gas through the lightpipe.

      8.7   Beamsplitter - If  the  spectrometer is thermostated,  maintain the
beamsplitter at  a temperature slightly  above  ambient at all  times.    If the
spectrometer is  not thermostated,  minimize exposure  of the beamsplitter  to
atmospheric water vapor.


9.0   METHOD PERFORMANCE

      9.1   Method 8410  has  been  in  use at the  U.S.  Environmental  Protection
Agency Environmental Monitoring  Systems Laboratory  for  more than  two years.
Portions of  it  have been reviewed by  key  members  of the FT-IR  spectroscopic
community (9).   Side-by-side comparisons  with  GC/MS  sample  analyses  indicate
similar demands  upon  analytical  personnel   for  the  two  techniques.   Extracts
previously subjected to GC/MS analysis are  generally compatible  with GC/FT-IR.
However,  it should be  kept in mind that  lightpipe windows are  typically water
soluble.   Thus,  extracts must be vigorously dried prior to  analysis.

      9.2   Table 4 provides  performance data for this method.
                                   8410 - 9                         Revision 0
                                                                September 1994

-------
10.0  REFERENCES

1.    Handbook  for  Analytical  Quality  Control   in   Water  and  Wastewater
      Laboratories;  U.S.  Environmental  Protection  Agency.    Environmental
      Monitoring and Support Laboratory.  ORD Publication Offices of Center for
      Environmental Research Information: Cincinnati, OH,  March  1979;  Sec.  4,
      EPA-600/4-79-019.

2.    Freeman, R.R.  Hewlett Packard Application  Note:   Quantitative Analysis
      Using a Purged Splitless  Injection Technique;  ANGC 7-76.

3.    Cole,  R.H.    Tables  of  Wavenumbers  for  the Calibration  of  Infrared
      Spectrometers; Pergamon:   New York, 1977.

4.    Grasselli, J.G.;  Griffiths, P.R.; Hannah,  R.W.   "Criteria for Presentation
      of Spectra from Computerized IR Instruments";  Appl.  Spectrosc. 1982, 36,
      87.

5.    Nyquist, R.A.  The Interpretation  of Vapor-Phase Infrared Spectra.  Group
      Frequency Data; Volume I. Sadtler Laboratories:  Philadelphia, PA, 1984.

6.    Socrates, G.   Infrared Characteristic Group Frequencies;  John Wiley and
      Sons:  New York,  NY, 1980.

7.    Bellamy, L.J.  The Infrared Spectra of Complex  Organic Molecules; 2nd ed.;
      John Wiley and Sons:  New York, NY, 1958.

8.    Szymanski, H.A.   Infrared Band Handbook,  Volumes  I  and II;  Plenum:  New
      York, NY, 1965.

9.    Gurka, D.F.  "Interim Protocol  for the Automated Analysis of Semivolatile
      Organic  Compounds   by   Gas   Chromatography/Fourier  Transform-Infrared
      Spectrometry"; Appl. Spectrosc. 1985,  39,  826.

10.   Griffiths, P.R.;   de  Haseth,  J.A.;  Azarraga, L.V.   "Capillary GC/FT-IR";
      Anal. Chem.  1983, 55, 1361A.

11.   Griffiths, P.P.;  de Haseth, J.A.  Fourier Transform-Infrared Spectrometry;
      Wiley-Interscience:  New York, NY, 1986.

12.   Gurka, D. F.; Farnham, I.;  Potter,  B.  B.;  Pyle, S.; Titus, R.  and Duncan,
      W.    "Quantitation    Capability    of     a   Directly    Linked    Gas
      Chromatography/Fourier Transform  Infrared/Mass Spectrometry System"; Anal.
      Chem., 1989, 61,   1584.
                                   8410 -  10                        Revision 0
                                                                September 1994

-------
                             TABLE 1.
FUSED SILICA CAPILLARY COLUMN GAS CHROMATOGRAPHIC/FOURIER TRANSFORM
    INFRARED  IDENTIFICATION LIMITS FOR BASE/NEUTRAL EXTRACTABLES
Compound
Acenaphthene
Acenaphthylene
Anthracene
Benzo(a)anthracene
Benzo(a)pyrene
Bis(Z-cnloroethyl) ether
Bis(2-chloroethoxy)methane
Bis(2-chloroisopropyl) ether
Butyl benzyl phthalate
4-Bromophenyl phenyl ether
2-Chloronaphthalene
4-Chloroaniline
4-Chlorophenyl phenyl ether
Chrysene
Di-n-butyl phthalate
Dibenzofuran
Diethyl phthalate
Dimethyl phthalate
Di-n-octyl phthalate
Di-n-propyl phthalate
1,2-Dichlorobenzene
1,3-Dichlorobenzene
1 ,4-Dichlorobenzene
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Bis-(2-ethylhexyl) phthalate
Fluoranthene
Fluorene
Hexachlorobenzene
Hexachl orocycl opentadi ene
Hexachloroethane
1 , 3 -Hexachl orobutadi ene
Isophorone
2-Methylnaphthalene
Naphthalene
Nitrobenzene
N-Nitrosodi methyl ami ne
N-Nitrosodi-n-propylamine
N-Nitrosodi phenyl ami ned
2-Nitroaniline
3-Nitroaniline
Identification
ng injected8
40(25)
50(50)
40(50)
(50)
(100)
70(10)
50(10)
50(10)
25(10)
40(5)
110
40
20(5)
(100)
20(5)
40
20(5)
20(5)
25(10)
25(5)
50
50
50
20
20
25(10)
100(50)
40(50)
40
120
50
120
40
110
40(25)
25
20(5)
50(5)
40
40
40
Limit
Mg/L"
20(12.5)
25(25)
20(25)
(25)
(50)
35(5)
25(5)
25(5)
12.5(5)
20(2.5)
55
20
10(2.5)
(50)
10(2.5)
20
10(2.5)
10(2.5)
12.5(5)
12.5(2.5)
25
25
25
10
10
12.5(5)
50(25)
20(25)
20
60
25
60
20
55
20(12.5)
12.5
10(2.5)
25(2.5)
20
20
20
»max, cm"
799
799
874
745
756
1115
1084
1088
1748
1238
851
1543
1242
757
1748
1192
1748
1751
1748
1748
1458
779
1474
1547
1551
1748
773
737
1346
814
783
853
1690
3069
779
1539
1483
1485
1501
1564
1583
                             8410 -  11
    Revision 0
September 1994

-------
                                   TABLE 1.
                                  (Continued)
                                 Identification Limit
      Compound                 ng injected8       M9/Lb            i>max, cm"1c
4-Nitroanil ine
Phenanthrene
Pyrene
1,2,4-Trichlorobenzene
40
50(50)
100(50)
50(25)
20
25(25)
50(25)
25(12.5)
1362
729
820
750
a  Determined using on-column  injection and the conditions of Sec. 7.3.  A medium
   band HgCdTe detector [3800-700 cm'1; D'value Upeak 1000 Hz, 1) 4.5 x  1010 cm
   Hz1/2W"1] type with  a  0.25 mm2 focal chip was used.  The GC/FT-IR  system is a
   1976 retrofitted model.   Values in parentheses  were determined with  a  new
   (1986)  GC/FT-IR system. A narrow band HgCdTe detector  [3800-750cm"1;  D'value
   (/Jpeak 1000 Hz, 1)  4 x 1010 cm Hz1'2^1] was  used.   Chromatographic conditions
   are those of Sec. 7.3.

b  Based on a 2 juL  injection  of a  one liter sample  that has been extracted  and
   concentrated to  a  volume of  1.0 mL.   Values in  parentheses were determined
   with a  new (1986) GC/FT-IR  system. A narrow band HgCdTe detector  [3800-750cm~
   1;  D'value  (/Ipeak 1000  Hz,  1) 4  x 1010 cm Hz1/2W"1]  was used.  Chromatographic
   conditions are those of Sec. 7.3.

0  Most intense IR peak and suggested quantitation  peak.

d  Detected as diphenylamine.
                                   8410 - 12                         Revision 0
                                                                 September 1994

-------
                                   TABLE 2.
      FUSED SILICA CAPILLARY COLUMN GAS CHROMATOGRAPHIC/FOURIER TRANSFORM
   INFRARED ON-LINE AUTOMATED IDENTIFICATION LIMITS FOR ACIDIC EXTRACTABLES
                                    Identification Limit
Compound
ng injected8
j>max, cm"
Benzoic acid
2-Chlorophenol
4-Chlorophenold
4-Chloro-3 -methyl phenol
2-Methyl phenol
4-Methyl phenol
2,4-Dichlorophenol
2,4-Dinitrophenol
4, 6-Dinitro-2-methyl phenol
2-NitrophenoTd
4-Nitrophenol
Pentachlorophenol
Phenol
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
70
50
100
25
50
50
50
60
60
40
50
50
70
120
120
35
25
50
12.5
25
25
25
30
30
20
25
25
35
60
60
1751
1485
1500
1177
748
1177
1481
1346
1346
1335
1350
1381
1184
1470
1458
*  Operating conditions are the same as those cited in Sec. 7.3.

b  Based on a 2 /LtL injection of a one liter sample that has been extracted and
   concentrated to a volume of 1.0 mL.

0  Most intense IR peak and suggested quantitation peak.

d  Subject to interference from co-eluting compounds.
                                   8410  -  13
                                     Revision 0
                                 September 1994

-------
         TABLE 3.
GAS-PHASE GROUP FREQUENCIES
Number of
Functionality Class Compounds
Ether





Ester


Nitro





Nitrile

Ketone


Amide
AT kyne
Acid




Phenol







Aryl, Alkyl
Benzyl , Al kyl
Di aryl
Dial kyl
Alkyl, Vinyl

Unsubstituted Aliphatic
Aromatic
Monosubstituted Acetate
Aliphatic



Aromatic

Aliphatic
Aromatic
Aliphatic (acyclic)
(a,|8 unsaturated)
Aromatic
Substituted Acetamides
Aliphatic
Aliphatic

Dimerized-Aliphatic
Aromatic

1,4-Disubstituted


1,3-Disubstituted


1,2-Disubstituted

14
3
5
12
3

29
11
34
5



18

9
9
13
2
16
8
8
24
22
2
10
10
15
15
15
10
10
10
6

Frequency
Range, i/cm"1
1215-1275
1103-1117
1238-1250
1084-1130
1204-1207
1128-1142
1748-1761
1703-1759
1753-1788
1566-1594
1548-1589
1377-1408
1327-1381
1535-1566
1335-1358
2240-2265
2234-2245
1726-1732
1638-1699
1701-1722
1710-1724
3323-3329
3574-3580
1770-1782
3586-3595
3574-3586
1757-1774
3645-3657
1233-1269
1171-1190
3643-3655
1256-1315
1157-1198
3582-3595
1255-1274
                                    (continued)
         8410 - 14
    Revision 0
September 1994

-------
TABLE 3.
(Continued)
Functional ity
Alcohol






Amine


Alkane



Aldehyde





Benzene





Class
Primary Aliphatic


Secondary Aliphatic

Tertiary Aliphatic

Primary Aromatic
Secondary Aromatic
Al iphatic




Aromatic


Aliphatic


Monosubstituted





Number of
Compounds
20
11
16
17
10
10
6
15
5
10
14



12
12
12
6
6
6
7
24
24
11
23
25
Frequency
Range, j>cm~1
3630-3680
1206-1270
1026-1094
3604-3665
1231-1270
3640-3670
1213-1245
3480-3532
3387-3480
760- 785
2930-2970
2851-2884
1450-1475
1355-1389
1703-1749
2820-2866
2720-2760
1742-1744
2802-2877
2698-2712
1707-1737
1582-1630
1470-1510
831- 893
735- 790
675- 698
8410 - 15
    Revision 0
September 1994

-------
    TABLE  4.   FUSED  SILICA CAPILLARY  COLUMN GC/FT-IR QUANTITATION RESULTS
Compound
Concentration
 Range, and
Identification
 Limit, nga
 Maximum
Absorbanceb
Correlation
Coefficient"1
 Integrated
Absorbance0
 Correlation
Coefficient11
Acenaphthene
Acenaphthylene
Anthracene
Benzo(a)anthracene
Benzoic acid
Benzo(a)pyrene
Bis(2-chloroethoxy)methane
Bis(2-chloroethyl) ether
Bis(2-chloroisopropyl) ether
4-Bromophenyl phenyl ether
Butyl benzyl phthalate
4-Chloroaniline
4-Chloro-3 -methyl phenol
2-Chloronaphthalene
2-Chlorophenol
4-Chlorophenole
4-Chlorophenyl phenyl ether
Chrysene
Dibenzofuran
Di-n-butyl phthalate
1,2-Dichlorobenzene
1,3'Dichlorobenzene
1,4-Dichlorobenzene
2,4-Dichlorophenol
Dimethyl phthalate
Dimethyl phthalate
Dinitro-2-methyl phenol
2,4-Dinitrophenol
2,4-Dinitrotoluene
2,6-Dinitrotoluene
Di-n-octyl phthalate
Bis(2-ethylhexyl) phthalate
Fluoranthene
Fluorene
Hexachl orobenzene
1,3-Hexachlorobutadiene
Hexachl orocycl opentadi ene
Hexachl oroethane
Isophorone
2-Methylnaphthalene
25-250
25-250
50-250
50-250
50-250
100-250
25-250
25-250
50-250
25-250
25-250
25-250
25-250
100-250
25-250

25-250
100-250
25-250
25-250
25-250
25-250
25-250
25-250
25-250
25-250
50-250
50-250
25-250
25-250
25-250
25-250
25-250
25-250
50-250
50-250
100-250
25-250
25-250
50-250
0.9995
0.9959
0.9969
0.9918
0.9864
0.9966
0.9992
0.9955
0.9981
0.9995
0.9999
0.9991
0.9975
0.9897
0.9976

0.9999
0.9985
0.9697
0.9998
0.9937
0.9985
0.9994
0.9964
0.9998
0.9998
0.9936
0.9920
0.9966
0.9947
0.9983
0.9991
0.9983
0.9987
0.9981
0.9960
0.9862
0.9986
0.9984
0.9981
0.9985
0.9985
0.9971
0.9921
0.9892
0.9074
0.9991
0.9992
0.9998
0.9996
0.9994
0.9965
0.9946
0.9988
0.9965

0.9997
0.9984
0.8579
0.9996
0.9947
0.9950
0.9994
0.9969
0.9996
0.9997
0.9967
0.9916
0.9928
0.9966
0.9991
0.9993
0.9966
0.9989
0.9995
0.9979
0.9845
0.9992
0.9990
0.9950
                                                                    (continued)
                                   8410 - 16
                                      Revision  0
                                  September  1994

-------
                             TABLE 4.   (Continued)
Compound
Concentration
  Range, and
Identification
  Limit, nga
  Maximum
Absorbanceb
Correlation
Coefficient11
 Integrated
Absorbance0
 Correlation
Coefficient11
2-Methyl phenol
4-Methyl phenol
Naphthalene
2-Nitroaniline
3-Nitroaniline
4-Nitroaniline
Nitrobenzene
2-Nitrophenole
4-Nitrophenol
N-Nitrosodi methyl ami ne
N-Nitrosodiphenylamine
N-Nitrosodi -n-propylamine
Pentachlorophenol
Phenanthrene
Phenol
Pyrene
1 , 2 , 4-Trichl orobenzene
2,4,5-Trichlorophenol
2,4,6-Trichlorophenol
25-250
25-250
25-250
25-250
25-250
25-250
25-250

50-250
25-250
25-250
25-250
50-250
25-250
25-250
50-250
50-250
25-250
25-250
0.9972
0.9972
0.9956
0.9996
0.9985
0.9936
0.9997

0.9951
0.9982
0.9994
0.9991
0.9859
0.9941
0.9978
0.9971
0.9969
0.9952
0.9969
0.9964
0.9959
0.9954
0.9994
0.9990
0.9992
0.9979

0.9953
0.9993
0.9971
0.9995
0.9883
0.9989
0.9966
0.9977
0.991
0.9966
0.9965
a  Lower end of range is at or near the identification limit.

b  FT-IR scan with highest absorbance plotted against concentration.

c  Integrated absorbance of  combined  FT-IR scans which occur  at  or above the
   chromatogram peak half-height.

d  Regression analysis  carried  out  at four concentration levels.   Each level
   analyzed in duplicate.  Chromatographic conditions are stated in Sec. 7.3.

e  Subject to interference from co-eluting compounds.
                                  8410  -  17
                                     Revision  0
                                 September 1994

-------
                                             METHOD  8410
               GAS  CHROMATOGRAPHY/FOURIER TRANSFORM  INFRARED  (GC/FT-IR)
               SPECTROMETRY  FOR SEMIVOLATILE  ORGANICS:   CAPILLARY  COLUMN
(   Start   j
  7.1 Sample
  preparation
   prior to
   GC/FT-IR
   analysis.
 7.2 Optional
     Gel
  Permeation
  Cleanup of
  extracts.
  7.3 Initial
 Calibration;
recommended
  GC/FT-IR
 conditions.
  7.4 Check
  detector
 centerburst
  intensity.
  7.5 Column
   Interface
  Sensitivity.
  7.6 Adjust
interferometer
   drive air
  pressure.
                        7.10.1 Analyze
                        extracts using
                         conditions of
                         Section 7.3.
                      7.10.2 GC/FT-IR
                       Identification;
                      compare analyte
                        IR spectrum;
                          report.
   7.7 MC
   Detector;
  centerburst
intensity  <75%
  plot max of
    Section
     7.4
                          7.10.3
                      Retention Time;
                      compare RRT of
                        analyte with
                         standard.
                        7.7 Replace
                          Source.
                       7.10.4 Report
                      compound class
                       if no library
                      match is found.
                       7.8 Frequency
                        Calibration.
                          7.10.5
                        Quantitation
                          desired.
                      7.9 Determine
                      min. identifiable
                       quantities of
                        analyte of
                         interest.
                                                                    7.10.6 Standard
                                                                    calibration curve
                                                                      of cone. vs.
                                                                     integrated IR
                                                                      absorbance.
                          7.10.6
                        Quantitation
                       by integrated
                        absorbance?
 7.9.1 Prepare
    plot of
 lightpipe T vs.
MCT centerburst
   intensity.
                                                                       7.10.8 Is
                                                                      GC Detector
                                                                    used in tandem
                                                                       with FT-IR
                                                                       detector?
                       7.10.7 Standard
                         calibration
                       curve of cone.
                       vs. max. IR band
                          intensity.

\ Yes

/
L*
7.10.8
Supplemental
GC Detector
Technique.


                                               8410  -  18
                                                                     Revision  0
                                                                September 1994

-------
4.4  MISCELLANEOUS SCREENING METHODS

         The following methods are included in this section:

         Method 3810:      Headspace
         Method 3820:      Hexadecane Extraction and Screening of Purgeable
                           Organics
         Method 4010:      Screening for Pentachlorophenol by Immunoassay
         Method 8275:      Thermal Chromatography/Mass Spectrometry (TC/MS)
                           for Screening Semi volatile Organic Compounds
                                 FOUR - 14                       Revision 2
                                                             September 1994

-------
                                 METHOD 3810

                                  HEADSPACE
1.0  SCOPE AND APPLICATION

     1.1  Method 3810 was formerly Method  5020  in the second edition of this
manual.

     1.2  Method 3810 is a static  headspace technique for extracting volatile
organic compounds from samples.    It  is  a  simple  method that allows large
numbers of samples to be screened in a relatively short period of time.  It is
ideal  for  screening  samples  prior  to  using  the  purge-and-trap  method.
Detection limits for this method may  vary widely among samples because of the
large variability and complicated matrices of waste samples.  The method works
best for compounds with boiling points of less than 125*C.  The sensitivity of
this method will depend on the equilibria of the various compounds between the
vapor and dissolved phases.

     1.3  Due to the variability of this method, this procedure is recommended
for use only as a  screening  procedure for other, more accurate determinative
methods (Methods 8010, 8015, 8020, 8030, and 8240).


2.0  SUMMARY OF METHOD

     2.1  The sample is collected  in  sealed  glass containers and allowed to
equilibrate at 90*C.  A sample of  the  headspace gas is withdrawn with a gas-
tight syringe for screening analysis using  the conditions specified in one of
the GC or GC/MS determinative methods (8010, 8015, 8020, 8030, or 8240).


3.0  INTERFERENCES

     3.1  Samples  can  be  contaminated  by  diffusion  of  volatile organics
(particularly chlorofluorocarbons and  methylene  chloride) through the sample
container septum during shipment and  storage.   A field sample blank prepared
from reagent water and  carried  through  sampling  and subsequent storage and
handling can serve as a check on such contamination.

     3.2  Contamination by carryover  can  occur  whenever high-level and low-
level samples are  sequentially  analyzed.    To  reduce carryover, the sample
syringe must be rinsed out  between  samples  with reagent water.  Whenever an
unusually concentrated sample  is  encountered,   it   should  be followed by an
analysis of reagent water.  It may  be  necessary to  wash out  the syringe with
detergent, rinse  with  distilled  water,  and  dry   in  a  105*C oven between
analyses.

     3.3  Before processing any  samples,  the analyst  should demonstrate daily
through the analysis of an organic-free water or  solvent blank that the entire
analytical system  is  interference-free.


                                   3810 -  1
                                                          Revision      0
                                                          Date  September 1986

-------
4.0  APPARATUS AND MATERIALS

     4.1  Refer to the specific determinative method for appropriate apparatus
and materials.

     4.2  Vials;    125-mL  Hypo-Vials   (Pierce   Chemical  Co.,  #12995,  or
equivalent), four each.

     4.3  Septa:  Tuf-Bond  (Pierce #12720 or equivalent).

     4'4  Seals;  Aluminum  (Pierce #132141 or equivalent).

     4.5  Crimper;  Hand  (Pierce #13212 or equivalent).

     4«6  Syringe;  5-mL,   gas-tight  with  shutoff  valve and chromatographic
needles.

     4'7  Microsyringe:   250- or 500-uL.

     4.8  Water  bath;    Heated,  with  concentric  ring  cover,   capable  of
temperature controT~[+5*C). The bath should be  used in  a  hood.


5.0 REAGENTS

     5.1   Refer to   the   specific  determinative method  and  Method  8000  for
preparation of calibration  standards.


6.0 SAMPLE COLLECTION,  PRESERVATION, AND  HANDLING

     6.1   Refer  to  the  introductory   material   to   this  chapter,   Organic
Analytes,  Section 4.1.


 7.0  PROCEDURE

      7.1   Gas  chromatographic   conditions  and   Calibration;     Refer  to  the
 specific  determinative method  for  GC operating  conditions and to Method 8000,
 Section 7.4, for calibration procedures.

      7.2   Sample preparation;

           7.2.1  Place 10.0 g  of a  well-mixed  waste   sample  into each of  two
      separate 125-mL septum-seal  vials.

           7.2.2  Dose one sample vial  through the septum with  200 uL of a
      50 ng/uL  calibration  standard  containing  the   compounds  of interest.
      Label this "1-ppm spike."
                                   3810 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
         7.2.3  Dose  a  separate  (empty) 125-mL  septum  seal vial with 200 uL
    of the  same 50 ng/uL calibration standard.  Label this  "1-ppm standard."

         7.2.4  Place the  sample,  1-ppm-spike,  and 1-ppm-standard vials into
    a 90*C  water  bath for  1  hr.  Store the remaining sample vial at 4.0*C for
    possible  future analysis.

    7.3  Sample analysis:

         7.3.1  While maintaining  the vials   at  90°C,  withdraw  2 ml of the
    headspace gas with  a   gas-tight  syringe  and analyze by direct injection
    into a  GC.  The GC  should  be operated using the same GC conditions listed
    in the  method being screened (8010, 8015, 8020, 8030, or 8240).

         7.3.2  Analyze the  1-ppm  standard  and adjust  instrument sensitivity
    to give a minimum response of   at   least  2 times the background.  Record
    retention times  (RT)  and peak  areas of compounds of interest.

         7.3.3  Analyze the  1-ppm  spiked  sample   in  the same  manner.  Record
    RTs  and peak  areas.

         7.3.4  Analyze the  undosed sample as in  Paragraph  7.3.3.

         7.3.5  Use  the results obtained  to   determine  if the  sample  requires
    dilution or methanolic extraction  as  indicated  in Method 5030.
8.0  QUALITY CONTROL

     8.1  Before  processing  any  samples,  the  analyst  should  demonstrate
through the analysis of a distilled  water method blank that all  glassware and
reagents are interference-free.  Each  time  a  set of samples is extracted or
there is a  change  in  reagents,  a  method  blank  should  be processed as a
safeguard against chronic laboratory contamination.   The blank samples should
be carried through all stages of the sample preparation and measurement.

     8.2  Standard  quality  assurance  practices  should  be  used  with this
method.  Fortified samples  should  be  carried  through  all stages of sample
preparation  and  measurement;  they  should   be  analyzed  to  validate  the
sensitivity and accuracy of the analysis.    If the fortified waste samples do
not indicate sufficient sensitivity to detect less  than or equal to 1 ug/g of
sample, then the sensitivity of the instrument should be increased.


9.0  METHOD PERFORMANCE

     9.1  No data provided.
                                  3810 - 3
                                                         Revision      0
                                                         Date  September 1986

-------
10.0  REFERENCES

1.  Hachenberg, H. and  A.  Schmidt,  Gas  Chromatographlc Headspace Analysis,
Philadelphia:  Hayden & Sons Inc., 1979.

2.  Friant, S.L. and I.H.  Suffet,  "Interactive  Effects of Temperature, Salt
Concentration and  pH  on  Headspace  Analysis  for  Isolating  Volatile Trace
Organlcs in Aqueous Environmental Samples," Anal. Chem. 51, 2167-2172, 1979.
                                   3810 - 4
                                                          Revision
                                                          Date  September 1986

-------
     METHOO  3B1O

  HEAOSPACE  METHOD
  C
Start
    7. 1
        Set  GC
      opcrat ing
      cono1tIons
    7.2
   Prepare sample
    7.3
         Analyzc
        by direct
        inj ect ion
        into  a GC
7

3.5
di
fl
c
Determine
if cample
required
lution or
letnanol ic
xtractlon
  f     Stop       J
3810 - 5
                          Revision       0
                          Date  September 1986

-------
                                 METHOD 3820

          HEXADECANE EXTRACTION AND SCREENING OF PUR6EABLE ORGANICS
1.0  SCOPE AND APPLICATION

     1.1  This method is a screening  procedure for use with purge-and-trap GC
or GC/MS.  The results of this  analysis are purely qualitative and should not
be used as an alternative to more detailed and accurate quantitation methods.


2.0  SUMMARY OF METHOD

     2.1  An aliquot of sample is  extracted with hexadecane and then analyzed
by GC/FID.  The  results  of  this  analysis  will indicate whether the sample
requires dilution or methanolic extraction prior to purge-and-trap GC or GC/MS
analysis.


3.0  INTERFERENCES

     3.1  Method interferences  may  be  caused  by  contaminants in solvents,
reagents, and glassware.  All  these  materials must be routinely demonstrated
to be free from  contaminants  by  running  laboratory reagent blanks.  Matrix
interferences may be  caused  by  contaminants  that  are coextracted from the
sample.  The extent of matrix interferences will vary considerably from sample
to sample depending upon the nature and diversity of the water being sampled.

     3.2  The flame  ionization  detector  varies  considerably in sensitivity
when comparing aromatics and  halogenated  methanes and ethanes.  Halomethanes
are  approximately  20x   less   sensitive   than  aromatics  and  haloethanes
approximately  lOx  less   sensitive.     Low-molecular-weight,  water-soluble
solvents  (e.g., alcohols and  ketones)  will  not  extract  from the water, and
therefore will not be detected by GC/FID.


4.0  APPARATUS AND MATERIALS

     4.1  Balance;  Analytical, capable of accurately weighing 0.0001 gm.

     4.2  Gas  Chromatograph;    An   analytical   system   complete  with  gas
chromatograph suitable for  on-column  injection  and all required accessories
including  syringes,  analytical  columns,  gases,  detector,  and strip-chart
recorder  (or equivalent).   A  data  system  is recommended for measuring peak
heights and/or peak areas.

          4.2.1  Detector:    Flame  ionization  (FID).

          4.2.2  GC  column:    3-m  x  2-mm   I.D.  glass   column  packed with
     10%  OV-101  on  100/120  mesh  Chromosorb  W-HP   (or   equivalent).   The
     column temperature  should be  programmed  from  80*C to 280'C at 16*C/min
     and  held at 280*C  for  10 min.
                                   3820 - 1
                                                          Revision       0
                                                          Date   September  1986

-------
     4.3  Centrifuge;   Capable of accommodating  50-mL glass  tubes.

     4.4  Vials and caps;  2-mL for GC autosampler.

     4.5  Volumetric flasks;    10-  and  50-mL   with  ground-glass stopper or
Teflon-lined screw-cap.

     4.6  Centrifuge tubes;  50-mL  with  ground-glass stopper or Teflon-lined
screw-cap.

     4.7  Pasteur plpets:  Disposable.

     4.8  Bottles;  Teflon-sealed screw-cap.


5.0  REAGENTS

     5.1  Hexadecane and methanol;  Pesticide quality or equivalent.

     5.2  Reagent water;    Reagent  water   is  defined  as  water  in which an
Interference is not observed  at   the  method detection limit of each parameter
of Interest.

     5.3  Stock standard solutions  (1.00 ug/uL):  Stock standard solutions can
be purchased as certifiedsolutions  or  can  be  prepared from pure standard
materials.

          5.3.1  Prepare stock standard solutions by  accurately weighing about
     0.0100 grams  of pure  material.    Dissolve  the  material  1n methanol  1n  a
     10-mL  volumetric  flask and dilute  to  volume (larger volumes may be  used
     at the convenience  of the analyst).    If compound purity is  certified at
     96%  or greater, the weight  can  be  used without correction  to calculate
     the  concentration of  the  stock  standard.   Commercially available stock
     standards may be  used if they are certified by  the manufacturer.

          5.3.2  Transfer   the  stock  standard  solutions  into Teflon-sealed
     screw-cap bottles.  Store at 4*C and protect from light.  These standards
     should be checked frequently for signs of degradation or evaporation.

     5.4  Standard mixture II;   Standard   mixture   #1 should contain benzene,
 toluene,  ethyl benzene,  and xylene.   Prepare a  stock solution containing these
 compounds as described In  Paragraph  5.3  and   then  prepare a working standard
 (through  dilution)  in  which the concentration of each compound in  the standard
 1s  100 ng/uL  1n methanol.

     5.5  Standard mixture 12;   Standard   mixture   #2 should contain n-nonane
 and  n-dodecane.     Prepare a  stock  solution   containing  these  compounds as
 described 1n  Paragraph 5.3.  Dilute  the  stock  standard with methanol so  that
 the  concentration  of each  compound is 100 ng/uL.
                                   3820 - 2
                                                          Revision      0
                                                          Date  September 1986

-------
6.0  SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

     6.1  See the introductory  material   to  this  chapter,  Organic Analytes,
Section 4.1.
7.0  PROCEDURE

     7.1  Sample preparation;

          7.1.1  Water;

               7.1.1.1  Allow the contents of the 40-mL sample vial  to come to
          room temperature.  Quickly transfer  the  contents of the 40-mL vial
          to a 50-mL volumetric flask.   Immediately add 2.0 mL of hexadecane,
          cap the flask, and shake  the  contents  vigorously  for 1 min.  Let
          phases separate.  Open the flask and add sufficient reagent water to
          bring the hexadecane layer into the neck of the flask.

               7.1.1.2  Transfer approximately 1 mL of the hexadecane layer to
          a 2.0-mL GC vial.    If  an  emulsion  is  present after shaking the
          sample, break it by:

                    1.  pulling the emulsion  through  a  small  plug of Pyrex
                        glass wool packed in a pipet, or

                    2.  transferring the  emulsion  to  a  centrifuge tube and
                        centrifuging for several min.

          7.1.2  Standards;

               7.1.2.1  Add  200 uL of the  working standard mixtures #1 and #2
          to   separate  40-mL  portions   of   reagent   water.     Follow  the
          instructions  in  Sections  7.1.1.1  and  7.1.1.2  'with the immediate
          addition of 2.0  mL of hexadecane.

          7.1.3  Sediment/Soil;

               7.1.3.1  Add  approximately 10 g of sample (wet weight) to 40 mL
          of  reagent  water  in  a  50-mL  centrifuge  tube.    Cap  and shake
          vigorously for  1   min.    Centrifuge   the  sample  briefly.  Quickly
          transfer the  supernatant water to a 50-mL volumetric  flask.

               7.1.3.2  Follow the instructions  given  in Sections  7.1.1.1 and
          7.1.1.2, starting  with the addition of 2.0 mL of hexadecane.

      7.2  Analysis;

          7.2.1  Calibration:

               7.2.1.1  External standard  calibration;    The  GC/FID must be
          calibrated each  12-hour shiftforhalf of full-scale response when
          injecting 1-5   uL  of  each  extracted standard  mixture  #1 and #2
           (Paragraphs 5.4  and 5.5).
                                   3820 - 3
                                                          Revision      0
                                                          Date   September  1986

-------
     7.2.2  GC/FID  analysis:    Inject  the  same  volume  of hexadecane
extract for the sample  under  Investigation  as  was used to perform the
external standard calibration.  The  GC conditions used for the standards
analysis must also be the same as those used to analyze the samples.

     7.2.3  Interpretation of the  GC/FID  chromatograms:    There are two
options for interpretation of the GC/FID results.

          7.2.3.1  Option  A;    The  standard  mixture  #1  is  used  to
     calculate an  approximate  concentration  of  the  aromatics  in the
     sample.  Use this information  to  determine the proper dilution for
     purge-and-trap if the  sample  is  a  water.    If  the  sample is a
     sediment/soil, use this information  to determine which GC/MS purge-
     and-trap method (low- or high-level) should  be used.   If aromatics
     are absent from the sample  or  obscured by higher concentrations of
     other purgeables, use Option B.

          7.2.3.2  Option B;  The response of standard mixture #2 is used
     to  determine  which  purge-and-trap   method  should  be  used  for
     analyzing a sample.  All purgeables of interest have retention times
     less  than  the  n-dodecane  retention  time.    A  dilution  factor
     (Paragraph 7.2.4.1.3) may be calculated  for water samples, and an X
     factor  (Paragraph 7.2.4.2.3) for soil/sediment samples, to determine
     whether the low-  or  high-level  purge-and-trap procedure should be
     used.

     7.2.4  Analytical decision point;

          7.2.4.1  Water  samples;   Compare  the  hexadecane sample extract
     chromatograms against an extracted  standard chromatogram.

                7.2.4.1.1   If no peaks   are  noted,   analyze   a  5-mL water
          sample by  the purge-and-trap  method.

                7.2.4.1.2   If peaks  are   present   prior to  the  n-dodecane
          peak  and   aromatics   are  distinguishable,  follow   Option   A
           (Paragraph 7.2.3.1).

                7.2.4.1.3   If peaks  are   present   prior to  the  n-dodecane
          but  the  aromatics   are   absent or   indistinguishable,  Option  B
          should be  used  as  follows:    If all  peaks (prior to n-dodecane)
          are  <3%  of the  n-nonane,   analyze  5  ml of water sample  by the
          purge-and-trap  method.   If  any   peak   is  >3%  of the n-nonane,
          measure  the area of the  major peak and calculate the necessary
          dilution factor as follows:

            dilution  factor = 50  x area  of major peak in sample
                                       peak  area  of n-nonane

          The  water sample should be  diluted using the calculated  factor
          just prior to purge-and-trap  GC or GC/MS analysis.
                              3820 - 4
                                                     Revision      0
                                                     Date  September 1986

-------
               7.2.4.2   Soil/sediment  samples:   Compare the hexadecane sample
          extract  chromatograms  against  an  extracted standard chromatogram.

                    7.2.4.2.1   If no peaks  are  noted,  analyze  a  5-g  sample by
               the low-level purge-and-trap procedure.

                    7.2.4.2.2   If peaks  are  present  prior  to the n-dodecane
               and aromatics are  distinguishable,  follow  Option A  using the
               concentration information given  in Table 1 to determine whether
               to  analyze the  sample   by a  low- or high-level purge-and-trap
               technique.

                    7.2.4.2.3   If peaks  are  present  prior  to n-dodecane but
               aromatics  are   absent  or  indistinguishable,   use  Option  B.
               Calculate an X  factor  for  the  sample  using  the  following
               equation:

                      X factor = area  of major  peak in sample
                                      area  of n-nonane

               Use the  information provided  in Table  1 to determine how the
               sample should be handled  for GC/MS analysis.

                    7.2.4.2.4   If  a   high-level   method   is   indicated,  the
               information provided in Table 2 can  be used to  determine the
               volume of methanol extract to add  to 5 ml of reagent  water for
               analysis (see Methods  5030  and  8240 for methanolic extraction
               procedure).


8.0  QUALITY CONTROL

     8.1  It is recommended that a reagent  blank be  analyzed by this  screening
procedure to ensure that no  laboratory   contamination  exists.   A blank  should
be performed for each set of samples undergoing extraction  and  screening.


9.0  METHOD PERFORMANCE

     9.1  No data available.
10.0  REFERENCES

1.  U.S. EPA  Contract  Laboratory  Program,  Statement  of  Work  for Organic
Analysis, July 1985, Revision.
                                  3820 - 5
                                                         Revision      0
                                                         Date  September 1986

-------
TABLE 1.  DETERMINATION OF GC/MS PURGE-AND-TRAP METHOD

                             Approximate
   X Factor              Concentration Range a                Analyze by

    0-1.0                  0-1,000 ug/kg                   Low-level method
    >1.0                   >1,000 ug/kg                    High-level method


      a This concentration range 1s based upon the response of aromatics to
GC/FID.  The concentration for halomethanes 1s 20x higher, and haloethanes
lOx higher, when comparing GC/FID responses.
 TABLE  2.   QUANTITY  OF  METHANOL  EXTRACT  REQUIRED  FOR ANALYSIS OF HIGH-LEVEL
       SOIL/SEDIMENTS

                              Approximate                       Volume  of
    X  Factor              Concentration  Range  a             Methanol  Extract  b

    0.25-5.0                 500-10,000  ug/kg                   100  ul
    0.5-10.0               1,000-20,000  ug/kg                   50  ul
    2.5-50.0               5,000-100,000 ug/kg                 10  ul
    12.5-250              25,000-500,000 ug/kg                 100  uL  of
                                                            1/50 dilution c


       a Actual concentration ranges could be   10  to  20 times  higher  than this
 1f the compounds are halogenated and the estimates are from GC/FID.

       b The volume  of methanol  added to  5 mL of water being  purged  should be
 100 uL.  Therefore 1f the amount of methanol  extract  required  is  less than 100
 uL, additional  methanol  should  be  added  to  maintain  the constant  100-uL
 volume.

       c Dilute an aliquot of  the  methanol  extract   and then take 100  uL for
 analysis.
                                   3820 - 6
                                                          Revision      0
                                                          Date  September 1986

-------
                     METHOD 3820

HEXAOECANE EXTRACTION AND SCREENING OF PURGEABLE ORGANICS

7. 1



Prepare sample

7.3.1



Calibrate
6C/FIO each
13-hour shift

7.2.3



Perform
GC/FIO analysis
                      o
              3820 - 7
                                      Revision       0
                                      Date  September 1986

-------
                                       METHOD 3620

                HEXAOECANE EXTRACTION AND SCREENING OF  PURGEABLE  ORGANICS
                                       (Continued!
7 .2 .A
Compare chroma togr-ams
of nexadecane sample
extract and extracted
      standard
     Are aromatic
   distinguishable?
                                                    7.2.4
                          Compare chromatograms
                          of hexadecane sample
                          extract and extracted
                                standard
                             7.2.4.2
                              Use low-level
                              purge-ana—trap
                                procedure
    Use standard
mixture 
-------
                                  METHOD 4010

                SCREENING FOR PENTACHLOROPHENOL BY IMMUNOASSAY
1.0  SCOPE AND APPLICATION

      1.1  Method  4010 is  a procedure  for screening  solids such  as soils,
sludges,   and  aqueous  media   such  as   waste   water  and   leachates  for
pentachlorophenol  (PCP) (CAS Registry 87-86-5).

      1.2  Method 4010 is recommended for screening samples to determine whether
PCP  is  likely to be present  at  concentrations above 0.5 mg/Kg  for solids or
0.005 mg/L  for aqueous samples.   Method  4010 provides  an estimate  for the
concentration of PCP by comparison with a standard.

      1.3  Using the test  kits  from which this method  was  developed,  95 % of
aqueous samples containing 2 ppb or less  of  PCPs will produce a negative result
in the 5 ppb test configuration.   Also, 95 % of soil samples  containing 125 ppb
or less of PCBs will produce a negative result in the 500 ppb test configuration.

      1.4  In cases where the exact concentration of PCP is required, additional
techniques (i.e.,  gas  chromatography (Method 8040) or gas chromatography/mass
spectrometry  (Method 8270)) should be used.


2.0  SUMMARY OF METHOD

      2.1    Test  kits are  commercially  available  for this  method.    The
manufacturer's  directions  should  be followed.    In  general,  the method  is
performed using a water sample or  an extract of  a soil  sample.   Sample and an
enzyme conjugate reagent are added to  immobilized antibody. The enzyme conjugate
"competes" with PCP present  in the sample for binding to immobilized anti-PCP
antibody.  The test is  interpreted by comparing the response produced by testing
a sample to the response produced by testing standard(s) simultaneously.


3.0  INTERFERENCES

      3.1  Compounds that are chemically similar may cause a positive test (false
positive) for PCP.  The test kit used  in preparation of this method was evaluated
for interferences.  Table 1  provides the concentration of compounds  found to give
a false positive test at the indicated concentration.

      3.2  Other compounds  have been tested for cross reactivity with PCP, and
have been demonstrated to not interfere with the specific kit tested.  Consult
the  information  provided  by the manufacturer  of  the kit used  for additional
information regarding cross reactivity with  other compounds.

      3.3  Storage and  use temperatures may modify the method performance. Follow
the manufacturer's directions for storage and use.
                                    4010-1                          Revision 0
                                                                   August 1993

-------
4.0  APPARATUS AND MATERIALS

      4.1  PENTA RISc Test Kits  (EnSys, Inc.), or equivalent.  Each commercially
available test kit will  supply or specify the apparatus and materials necessary
for successful completion of the test.


5.0  REAGENTS

      5.1   Each commercially  available  test kit  will  supply or  specify the
reagents necessary for successful completion of the test.


6.0  SAMPLE COLLECTION, PRESERVATION, AND HANDLING

      6.1  See  the  introductory material  to  this chapter,  Organic Analytes,
Section 4.1.
7.0  PROCEDURE

      7.1  Follow the manufacturer's instructions for the test kit being used.
Those test kits used must meet or exceed the performance indicated  in Tables 2-3.


8.0  QUALITY CONTROL

      8.1  Follow the manufacturer's instructions for the test kit  being used for
quality control procedures specific to the test kit used. Additionally, guidance
provided in Chapter One should be followed.

      8.2   Use  of  replicate  analyses,  particularly  when results  indicate
concentrations near  the action  level,   is  recommended to  refine  information
gathered with the kit.

      8.3  Do not use test kits past their expiration date.

      8.4  Do not use tubes or reagents designated for use with other kits.

      8.5   Use  the test kits  within their specified  storage  temperature and
operating temperature limits.

      8.6  Method 4010 is intended for field or laboratory  use.  The appropriate
level of quality assurance should  accompany  the application of this method to
document data quality.


9.0  METHOD PERFORMANCE

      9.1  This  method  has  been applied  to  a  series  of  groundwater,  process
water, and  wastewater samples from industries  which  use  PCP,  and  the results
compared with  GC/MS  determination  of  PCP (Method 8270).   These  results are
provided in Table 2.  These results represent determinations by two laboratories.


                                    4010-2                          Revision 0
                                                                   August 1993

-------
      9.2  This method  has  been applied to a  series  of soils from industries
which use PCP and the results compared with GC/MS determination of PCP via Method
8270.    These results  are  provided in  Table 3.    These  results  represent
determinations by two laboratories.
10.0 REFERENCES

1.   J.P. Mapes, K.D.  McKenzie, L.R. McClelland, S.  Movassaghi,  R.A. Reddy, R.L.
     Allen,   and  S.B.   Friedman,   "Rapid,   On-Site   Screening   Test   for
     Pentachlorophenol in Soil and Water  -  PENTA-RISc™", Ensys Inc., Research
     Triangle Park, NC 27709

2.   J.P. Mapes, K.D.  McKenzie, L.R. McClelland, S.  Movassaghi,  R.A. Reddy, R.L.
     Allen,  and S.B.  Friedman,   "PENTA-RISc    -  An  On-Site  Immunoassay  for
     Pentachlorophenol in Soil",  Bull.  Environ.  Contam.  Toxicol., 49:334-341,
     1992.

3.   PENTA-RISc™ Instructions  for Use,  Ensys  Inc.
                                    4010-3
 Revision 0
August 1993

-------
Table 1
Cross Reactivity for PCPa
Compound
2,6-Dichlorophenol
2,4,6-Trichlorophenol
2,4,5-Trichlorophenol
2,3, 4-Tri chl orophenol
2,3,5,6-Tetrachlorophenol
Tetrachl orohydroqui none
Concentration (mg/Kg)
in Soil to Cause a
False Positive for PCP
at 0.5 mg/Kg
700
16
100
400
1.2
500
Concentration (ng/L)
in Water to Cause a
False Positive for PCP
at 5 /xg/L
600
100
500
600
7
>1500
a for PENTA RISc Test Kit (EnSys, Inc.)
                               4010-4
 Revision 0
August 1993

-------
Table 2
Comparison of Immunoassay* with GC/MS
Water Matrix
Sample Type
groundwater





process water


wastewater



run-oft
















Screening Results (ppm)
0.005


>



>
>



>

>
>






>








0.05


<



>
>

>

>

>
<
>
>
>

>
>
<

>






0.1

>




<
>
>
>

<
>
<












>



0.5

<




<
<
<
<


<

















1
>



>
>



<
>




<
<
<

<
<

>
<
>
>



>
5
>


>
<
<




<







>







<
3
>

50
<


<

























<
Concentration measured
by GC/MS
3.5
0.35
<0.1
8.2
2.8
2.9
0.21
0.17
0.12
0.6
1.4
<0.1
0.17
<0.1
0.034
0.098
O.OS4
0.086
2.1
0.073
0.026
0.006
0.169
0.239
0.190
0.114
0.346
1.1
19
4.3

Does screening test agree with
GC/MS determination?
no
yes
yes
yes
yes
yes
no
yes
yes
no
yes
yes
yes
yes
yes
yes
yes
yes
no
yes
no
yes
no
yes
no
no
yes
yes
yes
yes
    > - screening test Indicates that the sample concentration Is greater than the test concentration

    < > screening test Indicates that the sample concentration Is less than the test concentration

• for PENTA RISc Test Kit (EnSys, Inc.)
                                                                4010-5
  Revision  0
August   1993

-------
Table 3
Comparison of Immunoassay* with GC/MS
Soil Matrix
Screening Results (ppm) |
0.5
>
>
<
<
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
>
<
<
<
>
<
>
>
<
>
>
5
>
>
<
<
>
<
>
>
<
>
<
>
<
>
>
>
>
>
>
>
>
<
>
<
<
<
<
<
<
>
<
>
>
50
>
<
<
<
>
<
<
<
<
>
<
<
<
<
<
<
<
>
>
>
>
<
<
<
<
<
<
<
<
>
<
>
<
Concentration measured by GC/MS
1100
88
0.31
0.72
315
1.5
6.4
9
1.9
46
<1
21
3.3
4
11
18
33
54
65
74
63
1.1
14.3
<1
<1
<1
3.9
<1
1.4
46
<1
142
106
Does screening test agree with
GC/MS determination?
yes
no
yes
no
yes
yes
yes
yes
yes
no
yes
yes
yes
no
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
no
yes
yes
no
4010-6
 Revision 0
August 1993

-------
Table 3
Continued
Screening Results (ppm)
0.5
>
>
>
>
>
<
<
>
5
>
>
<
>
>
<
<
<
50
>
>
<
<
>
<
<
<

Concentration measured by GC/MS
117
56
2.5
3.5
143
nd
0.02
5
Does screening test agree with
GC/MS determination?
yes
yes
yes
no
yes
yes
yes
yes
> - screening test Indicates that the sample concentration Is greater than the test concentration

< - screening test Indicates that the sample concentration Is less than the test concentration

• tor PENTA RISc Test Kit (EnSys, Inc.)
                                       4010-7
  Revision  0
August  1993

-------
                                  METHOD  8275

             THERMAL CHROMATOGRAPHY/MASS SPECTROMETRY (TC/MS) FOR
                   SCREENING SEMIVOLATILE ORGANIC COMPOUNDS
1.0   SCOPE AND APPLICATION

      1.1   Method  8275  is  a  screening technique  that may  be  used  for  the
qualitative identification of semivolatile organic compounds in extracts prepared
from nonaqueous solid wastes and soils.  It  is not intended for use as a rigorous
quantitative method.   Direct  injection of a  sample may  be used  in  limited
applications.  The  following analytes  can  be  qualitatively determined  by this
method:
      Compound Name                                   CAS No.a


      2-Chlorophenol                                   95-57-8
      4-Methylphenol                                  106-44-5
      2,4-Dichlorophenol                              120-83-2
      Naphthalene                                      91-20-3
      4-Chloro-3-methylphenol                          59-50-7
      1-Chloronaphthalene                              90-13-1
      2,4-Dinitrotoluene                              121-14-2
      Fluorene                                         86-73-7
      Diphenylamine                                   122-39-4
      Hexachlorobenzene                               118-74-1
      Dibenzothiophene                                132-65-0
      Phenanthrene                                     85-01-8
      Carbazole                                        86-74-8
      Aldrin                                          309-00-2
      Pyrene                                          129-00-0
      Benzo(k)fluoranthene                            207-08-9
      Benzo(a)pyrene                                   50-32-8


      8   Chemical  Abstract Services  Registry  Number.

      1.2   Method  8275  can be used  to qualitatively identify  most neutral,
acidic,  and basic organic compounds that  can be thermally desorbed  from a sample,
and are  capable of being eluted without derivatization as  sharp peaks  from a gas
chromatographic fused-silica capillary column  coated with  a slightly  polar
silicone.

      1.3   This method  is  restricted to use by  or under the supervision  of
analysts experienced  in  the use  of gas chromatograph/mass  spectrometers  and
skilled  in  the  interpretation of mass spectra.  Each analyst must demonstrate the
ability  to generate acceptable results with this method.
                                   8275 - 1                         Revision 0
                                                                September 1994

-------
2.0   SUMMARY OF METHOD

      2.1   A portion of  the  sample  (0.010-0.100 g) is weighed  into  a sample
crucible.   The  crucible  is placed  in a pyrocell  and  heated.    The  compounds
desorbed from the sample are detected using a flame ionization  detector (FID).
The FID response is used  to calculate the  optimal  amount  of sample needed for
mass spectrometry.   A  second sample is desorbed and the compounds are condensed
on the head of a fused  silica capillary column.  The column is  heated using a
temperature program, and the effluent  from the column is  introduced into the mass
spectrometer.


3.0   INTERFERENCES

      3.1   Contamination by carryover can  occur whenever low-level samples are
analyzed after high-level  samples.  Whenever an  unusually concentrated sample is
encountered, it  should be followed by the analysis of an empty (clean) crucible
to check for cross contamination.
4.0   APPARATUS AND MATERIALS

      4.1   Thermal Chromatograph (TC) System

            4.1.1 Thermal chromatograph™,  Ruska Laboratories,  or equivalent.

            4.1.2 Column  -  30  m x  0.25  mm  ID  (or 0.32  mm  ID),  1 urn  film
      thickness, si 1icone-coated, fused-silica capillary column (J&W Scientific
      DB-5 or equivalent).

            4.1.3 Flame Ionization detector (FID).

      4.2   Mass Spectrometer (MS) system

            4.2.1 Mass Spectrometer  -  Capable of scanning from 35  to  500 amu
      every one second or less,  using 70 volts (nominal) electron energy in the
      electron impact ionization mode.

            4.2.2 TC/MS interface -  Any GC-to-MS  interface producing acceptable
      calibration data in the concentration range of interest may be used.

            4.2.3 Data  System  -  A  computer  must  be  interfaced  to the  mass
      spectrometer.  The data system must allow the continuous acquisition and
      storage on machine-readable media of  all  mass spectra obtained throughout
      the  duration  of the  chromatographic program.   The computer  must  have
      software that can search  any GC/MS data file for ions of a specific mass
      (or group of masses) and that  can plot such  ion abundances versus time or
      scan  number.    This   type of  plot  is  defined as   an  extracted  ion
      chromatogram  (EIC).    Software  must  also  be  available that  allows for
      integration of the abundances  in, and EIC between, specified time or scan-
      number 1imits.
                                   8275 - 2                         Revision 0
                                                                September 1994

-------
      4.3   Tools and equipment

            4.3.1 Fused quartz spatula.

            4.3.2 Fused quartz incinerator ladle.

            4.3.3 Metal forceps for sample crucible.

            4.3.4 Sample crucible storage dishes.

            4.3.5 Porous fused quartz sample crucibles with lids.

            4.3.6 Sample crucible cleaning incinerator.

            4.3.7 Cooling rack.

            4.3.8 Microbalance, 1 g  capacity,  0.000001 g sensitivity, Mettler
      Model M-3 or equivalent.

      4.4   Vials -  10 mL, glass with Teflon lined screw-caps or crimp tops.

      4.5   Volumetric flasks, Class A - 10 ml to 1000 ml.


5.0   REAGENTS

      5.1   Reagent grade inorganic chemicals shall be  used in all  tests.  Unless
otherwise  indicated,  it is  intended  that all  reagents  shall conform  to the
specifications of the Committee on Analytical  Reagents of the American Chemical
Society, where such specifications are available.

      5.2   Solvents

            5.2.1 Methanol, CH3OH -  Pesticide grade or equivalent.

            5.2.2 Acetone, CH3COCH3  -  Pesticide  grade  or  equivalent.

            5.2.3 Toluene, C6H5CH3 - Pesticide grade or equivalent.

            5.2.4 Methylene chloride, CH2C12  -  Pesticide  grade or equivalent.

            5.2.5 Carbon disulfide,  CS2 - Pesticide grade or equivalent.

            5.2.6 Hexane, C6H14 -  Pesticide grade  or equivalent.

            5.2.7 Other suitable solvents - Pesticide grade or equivalent.

      5.3   Stock Standard solutions - Standard solutions may be prepared from
pure standard materials or purchased as certified solutions.

            5.3.1 Prepare stock standard solutions by weighing about 0.01 g of
      pure material.   Dissolve the  material  in  pesticide  quality  acetone,  or
                                   8275 - 3                         Revision 0
                                                                September 1994

-------
      other suitable solvent,  and dilute to 10 ml in  a volumetric flask.  Larger
      volumes may be used at the convenience of the  analyst.

            5.3.2 Transfer the stock standard  solutions  into  glass vials with
      Teflon lined screw-caps or crimp tops.   Store at  -10°C  to  -20°C or less
      and  protect  from  light.   Stock  standard  solutions  should  be checked
      frequently for signs of degradation  or  evaporation,  especially prior to
      use in preparation of calibration standards.

            5.3.3 Stock  standard solutions  must be  replaced  after  1  year,  or
      sooner  if comparison  with  quality  control check  samples  indicates  a
      problem.

      5.4   Internal Standard solutions - The internal standards recommended are
l,4-dichlorobenzene-d4,   naphthalene-d8,    acenaphthene-d10,   phenanthrene-d10,
chrysene-d12> and perylene-d12.  Other compounds  may be used as internal standards
as long as the requirements given  in Sec. 7  are met.  Dissolve about 0.200 g of
each compound with  a small volume  of carbon  disulfide.   Transfer  to a  50 ml
volumetric flask and dilute to volume with methylene  chloride,  so that  the final
solvent is approximately 20/80 (V/V) carbon  disulfide/methylene chloride.  Most
of the  compounds  are also soluble  in  small volumes of  methanol,  acetone,  or
toluene, except for perylene-d12.   Prior  to each analysis,  deposit about  10 juL
of the  internal standard  onto  the  sample  in  the  crucible.   Store internal
standard solutions at 4°C or less  before,  and  between,  use.

      5.5   Calibration  standards  - Prepare calibration  standards  within the
working range of the TC/MS  system.  Each standard should contain each analyte or
interest (e.g.   some or all of the compounds  listed in Sec. 1.1  may be included).
Each aliquot of calibration  standard should be spiked  with internal standards
prior to analysis.   Stock solutions should be stored  at -10°C to -20°C and should
be freshly prepared  once  a year,  or  sooner if check standards indicate a problem.
The daily calibration standard should be prepared weekly, and stored at 4°C.


6.0   SAMPLE COLLECTION, PRESERVATION,  AND HANDLING

      6.1   See the  introductory  material  to  this  Chapter,  Organic Analytes,
Sec. 4.1.
7.0   PROCEDURE

      7.1   Crucible Preparation

            7.1.1 Turn  on  the  incinerator  and let  it heat  for at  least  10
      minutes.  The bore of the incinerator should be glowing red.

            7.1.2 Load the sample  crucible  and lid  into the incinerator ladle
      and  insert  into  the incinerator bore.   Leave in the  incinerator for 5
      minutes, then remove and place on the cooling rack.

            7.1.3 Allow the crucibles and lids to cool for five minutes before
      placing them in the storage dishes.


                                   8275 - 4                         Revision 0
                                                                September 1994

-------
      CAUTION:    Do not touch the crucibles with your fingers.  This can
                  result in  a  serious  burn,  as well  as contamination of
                  the crucible.   Always  handle the sample crucibles and
                  lids with  forceps and tools specified.

      7.1.4 All  sample  crucibles  and  lids  required  for the  number of
analyses planned should be cleaned and  placed in  the storage dishes ready
for use.

7.2   Sample Preparation and Loading

      7.2.1 The  analyst should  take  care  in   selecting  a  sample  for
analysis, since the sample size is generally limited to 0.100 g or less.
This  implies  that  the  sample should be mixed  as thoroughly as possible
before taking an aliquot.  Because the sample size is limited, the analyst
may wish to analyze several  aliquots for determination.

      7.2.2 The  sample  should  be mixed or ground such that  a  0.010 to
0.100 g  aliquot can be  removed.   Remove  one sample  crucible  from  the
storage dish and place  it on  the microbalance.  Establish the tare weight.
Remove the sample crucible from the balance with  the forceps and place it
on a clean surface.

      7.2.3 Load an amount of  sample  into  the sample  crucible using the
fused  quartz  spatula.    Place  the  assembly  on  the  microbalance  and
determine the  weight of  the  sample.   For severely contaminated samples,
less  than  0.010 g  will  suffice,  while 0.050-0.100 g  is  needed  for  low
concentrations of contaminants.  Place the crucible lid on the crucible;
the sample is  now ready for  analysis.

7.3   FID Analysis

      7.3.1 Load the sample  into  the TC.   Hold the sample at 30°C for 2
minutes  followed  by linear  temperature  programmed heating  to  260°C at
30°C/minute.    Follow the temperature program with an  isothermal  heating
period of 10 minutes at 260°C, followed  by cooling back to 30°C.  The total
analysis cycle time is 24.2 minutes

      7.3.2 Monitor the  FID  response in real time during analysis,  and
note  the  highest response in  millivolts  (mV).   Use this  information to
determine  the  proper  weight  of  sample  needed  for  combined  thermal
extraction/gas chromatography/mass spectrometry.

7.4   Thermal  Extraction/GC/MS

      7.4.1 Prepare a calibration curve using a clean  crucible and lid by
spiking the compounds of interest at five concentrations into the crucible
and applying the internal standards  to the crucible  lid.   Analyze these
standards and establish response factors at different concentrations.

      7.4.2 Weigh  out  the  amount  of  fresh  sample  that will  provide
approximately 1000  to 3000 mv response.  For example, if 0.010 g of sample
gives an FID response of 500 mv,  then  0.020  to  0.060  g (0.040 g  ± 50 %)


                             8275 - 5                         Revision 0
                                                          September 1994

-------
should be used.  If 0.100 g gives 8000 mv, then 0.025 g ± 50 % should be
used.

      7.4.3 After weighing out the sample into the crucible, deposit the
internal standards (10 juL)  onto  the  sample.   Load the crucible into the
pyrocell, using  the  same temperature program  in  Sec. 7.3.1.   Hold the
capillary at 5°C during this time to focus  the released semivolatiles (the
intermediate trap is  held at 330°C to  pass  all compounds onto the column).
Maintain the splitter zone at  310°C, and the GC/MS  transfer line at 285°C.
After the isothermal  heating period is complete, temperature program the
column from 5°C to 285°C  at 10°C/minute  and hold at 285°C for 5 minutes.
Acquire data during the entire run time.

      7.4.4 If the response for any quantitation ion exceeds the initial
calibration curve range  of  the TC/MS  system,  a smaller sample should be
analyzed.

7.5   Data Interpretation

      7.5.1 Qualitative Analysis

            7.5.1.1     The   qualitative   identification  of  compounds
      determined  by  this  method  is  based  on retention  time, and  on
      comparison of the sample mass spectrum,  after background correction,
      with  characteristic  ions   in  a  reference   mass  spectrum.    The
      reference mass spectrum must be generated by the laboratory using
      the conditions of  this  method.   The characteristic  ions from the
      reference mass  spectrum  are  defined to be the three  ions of greatest
      relative intensity, or any ions  over 30%  relative intensity if less
      than three  such  ions  occur in the  reference  spectrum.   Compounds
      should be identified as present when the criteria below are met.

                  7.5.1.1.1   The intensities of the characteristic ions
            of a compound maximize in  the  same  scan  or within one scan of
            each  other.   Selection of  a peak  by a data  system target
            compound  search  routine where  the search  is based on the
            presence  of  a  target chromatographic  peak  containing  ions
            specific  for  the target  compound at  a  compound-specific
            retention time will  be accepted as meeting this criterion.

                  7.5.1.1.2   The RRT of  the  sample component is within
            ± 0.06 RRT units of the RRT of the standard component.

                  7.5.1.1.3   The    relative   intensities    of    the
            characteristic  ions   agree   within  30%  of  the  relative
            intensities   of   these   ions   in   the  reference  spectrum.
            (Example:    For  an  ion  with  an  abundance  of  50% in the
            reference spectrum,  the corresponding abundance in a sample
            spectrum can  range between 20% and 80%.)

                  7.5.1.1.4   Structural isomers that produce very similar
            mass  spectra should  be identified  as  individual  isomers if
            they   have   sufficiently   different   GC   retention  times.


                             8275 - 6                         Revision 0
                                                          September 1994

-------
      Sufficient  GC  resolution is achieved  if the height  of the
      valley between two isomer peaks is less than 25%  of the sum of
      the  two  peak  heights.   Otherwise,  structural   isomers are
      identified as isomeric pairs.

            7.5.1.1.5   Identification  is  hampered  when  sample
      components  are not  resolved chromatographically and produce
      mass spectra  containing  ions contributing by more than one
      analyte.   When gas chromatographic peaks obviously represent
      more than one  sample  component  (i.e.,  a broadened peak with
      shoulder(s)  or  a  valley  between  two  or  more  maxima),
      appropriate  selection  of   analyte  spectra  and  background
      spectra is  important.   Examination  of extracted ion current
      profiles  of appropriate ions  can aid  in the  selection  of
      spectra,  and in qualitative  identification of compounds.  When
      analytes  coelute (i.e.,  only  one  chromatographic peak  is
      apparent), the identification criteria  can be met, but each
      analyte spectrum will contain extraneous ions contributed by
      the coeluting compound.

      7.5.1.2     For samples containing components not associated
with the calibration standards, a  library search may be made for the
purpose of tentative identification.  The necessity to perform this
type  of  identification will be determined  by the purpose  of the
analyses  being  conducted.     Computer  generated  library  search
routines   should   not   use  normalization  routines   that   would
misrepresent the  library  or unknown  spectra when compared to each
other.  For example, the RCRA permit or waste delisting requirements
may require the  reporting  of non-target analytes. Only after visual
comparison of sample spectra with  the nearest  library  searches will
the  mass  spectral  interpretation specialist  assign  a tentative
identification.  Guidelines  for making tentative identification are:

      (1)   Relative intensities  of  major  ions in the reference
spectrum (ions  > 10% of the  most abundant ion)  should  be present in
the sample spectrum.

      (2)   The relative intensities  of  the major ions should agree
within ± 20%.   (Example:   For an ion with an  abundance  of 50% in the
standard spectrum, the  corresponding sample  ion abundance must be
within 30 and 70%).

      (3)   Molecular ions present in the reference spectrum should
be present in the sample spectrum.

      (4)   Ions  present  in the  sample spectrum  but not  in the
reference  spectrum  should  be  reviewed for   possible  background
contamination or presence of coeluting compounds.

      (5)   Ions present  in the reference spectrum but  not  in the
sample spectrum should  be reviewed for possible subtraction from the
sample spectrum because of  background contamination or coeluting.
Data system library reduction  programs  can  sometimes  create these
discrepancies.

                       8275 - 7                         Revision 0
                                                    September 1994

-------
8.0   QUALITY CONTROL

      8.1   Refer to Chapter One  and  Method  8000  for specific quality control
procedures.


9.0   METHOD PERFORMANCE

      9.1   Table 1 presents method performance data, generated using spiked soil
samples.  Method performance data in an aqueous matrix are not available.


10.0  REFERENCES

1.    Zumberge, J.E., C. Sutton, R.D. Worden, T. Junk, T.R. Irvin, C.B. Henry,
      V. Shirley,  and E.B.  Overton,  "Determination of  Semi-Volatile Organic
      Pollutants in Soils by Thermal Chromatography-Mass Spectrometry (TC/MS):
      an Assessment for Field Analysis," in preparation.
                                   8275  - 8                         Revision 0
                                                                September 1994

-------
                                TABLE 1
                    METHOD PERFORMANCE, SOIL MATRIX
Analyte
2-Chlorophenol
4-Methyl phenol
2,4-Dichlorophenol
Naphthalene
4-Chloro-3-methyl -phenol
1-Chloronaphthalene
2,4-Dinitrotoluene
Fluorene
Diphenylamine
Hexachlorobenzene
Dibenzothiophene
Phenanthrene
Carbazole
Aldrin
Pyrene
Benzo ( k) f 1 uoranthene
Benzo(a)pyrene
Average
Clay
30
10
23
77
9
96
7
9
5
68
20
11
4
3
7
4
4
% Recovery"
Silt
22
77
20
120
12
103
10
25
6
64
35
31
8
19
19
9
8

Subsoil
2
7
26
63
9
70
10
19
6
80
50
40
9
15
20
11
11
Mean
Recovery
18
31
23
87
10
90
9
18
6
71
35
24
7
12
15
8
8
Percent theoretical  recovery  based upon linearity of injections  deposited on
the crucible lid (slope and y-intercept).  Average of 9 replicates  (-10 mg
soil spiked with 50 ppm of analyte);  3  different  instruments  at 3 different
laboratories.
                               8275  - 9                         Revision 0
                                                            September 1994

-------
                                   TABLE  2
                CHARACTERISTIC IONS FOR SEMIVOLATILE COMPOUNDS
Compound
Primary
Ion
Secondary
Ion(s)
2-Chlorophenol
4-Methylphenol
2,4-Dichlorophenol
Naphthalene
4-Chloro-3-methyl-phenol
1-Chloronaphthalene
2,4-Dinitrotoluene
Fluorene
Diphenylamine
Hexachlorobenzene
Phenanthrene
Aldrin
Pyrene
Benzo(k)f1uoranthene
Benzo(a)pyrene
  128
  107
  162
  128
  107
  162
  165
  166
  169
  284
  178
   66
  202
  252
  252
  64,130
 107,108,77,79,90
 164,98
 129,127
 144,142
 127,164
  63,89
 165,167
 168,167
 142,249
 179,176
 263,220
 200,203
 253,125
 253,125
                                   8275 - 10
                                Revision 0
                            September 1994

-------
                      METHOD  8275
 THERMAL CHROMATOGRAPHY/MASS SPECTROMETRY  (TC/MS)  FOR
       SCREENING SEMIVOLATILE ORGANIC COMPOUNDS
Start
7.1 Prepare
crucible
^
r
7.2.2
Establish
tare weight
of crucible.
1
f
7.2.3 Place
sample in
crucible; establish
weight.
^
r
7.3.1 FID
Analysis using
linear temp.
programmed
heating.



\r
7.3.2 Using
FID response,
determine
sample weight
for TE/GC/MS.




>fc
•

7.4.1 f
cahbr
cur
>
repare
ve.
f
7.4.2 Prepare
amount of
sample for
appropriate
FID response.
^
r
7.4.3 Weigh
sample into
crucible; use
temp, program
in Sec. 7.3.1 .
/7.4.4\
/ Is N.
/ quantitation N^
f ion > initial
N. calib. curve .
N. range of /
XTC/MS?/
] No
I

7.5.1
Qualitative
Identification.
1
*
.Yes


r



7.4.4 Use
smaller
sample.

                            Stop
                       8275 -  11
    Revision 0
September 1994

-------
                                  APPENDIX

                             COMPANY REFERENCES


The  following  listing  of frequently-used  addresses  is  provided  for  the
convenience of users of this manual.   No endorsement  is intended or implied.


Ace Glass Company
1342 N.W. Boulevard
P.O. Box 688
Vineland, NJ  08360
(609) 692-3333

Aldrlch Chemical Company
Department T
P.O. Box 355
Milwaukee, WI  53201

Alpha Products
5570 - T W. 70th Place
Chicago, IL  60638
(312) 586-9810

Barneby and Cheney Company
E. 8th Avenue and N. Cassidy Street
P.O. Box 2526
Columbus, OH  43219
(614) 258-9501

Bio - Rad Laboratories
2200 Wright Avenue
Richmond, CA  94804
(415) 234-4130

Burdick & Jackson Lab Inc.
1953 S. Harvey  Street
Muskegon, MO  49442

Calgon Corporation
P.O. Box  717
Pittsburgh, PA   15230
(412) 777-8000

Conostan  Division
Conoco Speciality Products,  Inc.
P.O. Box  1267
Ponca City, OK   74601
(405) 767-3456
                                 COMPANIES -  1
                                                          Revision       0
                                                          Date   September  1986

-------
Corning Glass Works
Houghton Park
Corning, NY  14830
(315) 974-9000

Dohrmann, Division of Xertex Corporation
3240 - T Scott Boulevard
Santa Clara, CA  95050
(408) 727-6000
(800) 538-7708

E. M. Laboratories, Inc.
500 Executive Boulevard
Elmsford, NY  10523

Fisher Scientific Co.
203 Fisher Building
Pittsburgh, PA  15219
(412) 562-8300

General  Electric Corporation
3135 Easton Turnpike
Fairfield, CT  06431
(203) 373-2211

Graham Manufactory Co.,  Inc.
20  Florence Avenue
Batavia, NY   14020
(716) 343-2216

Hamilton Industries
1316  18th  Street
Two Rivers, WI  54241
(414) 793-1121

ICN Life Sciences  Group
3300  Hyland Avenue
Costa Mesa, CA  92626

Johns -  Manville  Corporation
P.O.  Box 5108
Denver,  CO 80217

Kontes  Glass  Company
8000 Spruce Street
Vineland,  NJ   08360

Milllpore  Corporation
80  Ashby Road
Bedford, MA  01730
 (617) 275-9200
 (800) 225-1380
                                 COMPANIES - 2
                                                          Revision
                                                          Date   September 1986

-------
National Bureau of Standards
U.S. Department of Commerce
Washington, DC  20234
(202) 921-1000

Pierce Chemical Company
Box 117
Rockford, IL  61105
(815) 968-0747

Scientific Glass and  Instrument,  Inc.
7246 - T Wynnwood
P.O. Box 6
Houston, TX  77001
(713) 868-1481

Scientific Products Company
1430 Waukegon Road
McGaw Park,  IL  60085
(312) 689-8410

Spex Industries
3880 - T and Park Avenue
Edison, NJ   08820