07/21/2003 10:23 FAX                                                                [?1006
             ::                    906R91102                                 "
                                          METHOD 245.6
                               DETERMINATION OF MERCURY IN TISSUES
                          BY COLO VAPOR ATOMIC ABSORPTION SPECTROMETRY
                        Edited by Larry B. Lobring and Billy B. Potter
                                  Inorganic Chemistry Branch
                                  Chemistry Research Division
                                        Revision 2.3
                                         April  1991
      frt
      C)
                        ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
                            OFFICE OF RESEARCH AND DEVELOPMENT
                           U.S.  ENVIRONMENTAL PROTECTION AGENCY
                                 CINCINNATI, OHIO  45268
                                           281       HEADQUARTERS LIBRARY
                                                     ENVIRONMENTAL PROTECTIOM
                                                     WASHINGTON, D.C ^f-"

-------
07/21/2003 10:29 FAX                                                               @]007/011
                                          METHOD 245.6

                               DETERMINATION OF MERCURY IN TISSUES
                          BY COLD VAPOR ATOMIC ABSORPTION SPECTROMETRY


          1-   SCOPE AND APPLICATION

              1.1  This procedure measures total mercury  (organic + inorganic) in
                   biological tissue samples.

              1.2  The range of the method is 0.2 to 5 M9/9-  The range may  be extend-
                   ed above or below the normal range by  increasing or decreasing
                   sample size or by optimizing instrument sensitivity.

          2.   SUMMARY OF METHOD

              2,1  A weighed portion of the  tissue sample is digested with sulfuric and
                   nitric acid at 58°C followed by overnight oxidation with  potassium
                   permanganate and potassium persulfate  at room temperature.  Mercury
                   in the digested sample is reduced with stannous chloride  to
                   elemental mercury and measured by the  conventional cold vapor atomic
                   absorption technique.

          3.   DEFINITIONS

              3.1  BIOCHEMICAL OXYGEN DEMAND (BOD) BOTTLE - BOD bottle, 300  ± 2 mL with
                   a ground glass stopper or an equivalent flask, fitted with a ground
                   .glass stopper.

              3.2  CALIBRATION BLANK - A volume of ASTM type II reagent water prepared
                   in the same manner (acidified) as the calibration standard.

              3.3  CALIBRATION STANDARD (CAL) - A solution prepared from the mercury
                   stock standard solution used to calibrate the instrument  response
                   with respect to analyte concentration.

              3.4  INSTRUMENT DETECTION LIMIT (IDL) - The mercury concentration that
                   produces a signal equal to three times the standard deviation of the
                   blank signal.

              3.5  LABORATORY FORTIFIED BLANK (LFB) - An aliquot of ASTM type II
                   reagent water to which known quantities of inorganic and/or organic
                   mercury are added in the laboratory.  The LFB is analyzed exactly
                   like a sample, and its purpose is to determine whether method
                   performance Is within accepted control limits.

              3.6  LABORATORY FORTIFIED SAMPLE MATRIX (LFM) - A portion of a tissue
                   sample to which known quantities of  calibration standard are added
                   in the laboratory.  The LFM is analyzed exactly like a sample, and
                   its purpose is to determine whether the sample matrix contributes


                                               282

-------
                                                                                  il007
07/21/2003 10:23 FAX
                    bias to the analytical  results.   The background concentrations of
                    the analytes in the sample matrix must be determined in a separate
                    aliquot and the measured values  in the LFM corrected for the
                    concentrations found,

               3.7  LABORATORY REAGENT BLANK (LRB)  - An aliquot of ASTM type II reagent
                    water that is treated  exactly as a sample including exposure to all
                    glassware, equipment,  and reagents used in analyses.  The LRB is
                    used to determine if method analyte or other interferences are
                    present in the laboratory environment, the reagents or apparatus.

               3.8  LINEAR DYNAMIC RANGE (LDR) - The concentration range over which the
                    analytical working curve remains linear.

               3.9  METHOD DETECTION LIMIT (MDL) -  The minimum concentration of mercury
                    that can be identified, measured and reported with 99% confidence
                    that the analyte concentration  is greater than zero and determined
                    from analysis of laboratory fortified tissue sample matrix (LFM).

               3.10 QUALITY CONTROL SAMPLE (QCS) -  A tissue sample containing known
                    concentration of mercury derived from externally prepared test
                    materials.  The QCS is obtained  from a source external to the
                    laboratory and is used to check laboratory performance.

               3.11 TISSUE SAMPLE - A biological sample matrix exposed to a marine,
                    brackish or fresh water environment.  It  is limited fay this method
                    to the edible tissue portion.

               3.12 STOCK STANDARD SOLUTION - A concentrated  solution containing mercury
                    prepared in the laboratory using assayed  mercuric chloride or stock
                    standard solution purchased from a reputable commercial source.

          4.    INTERFERENCES

               4.1  Interferences have been reported for waters containing sulfide,
                    chloride, copper and tellurium.   Organic  compounds which have broad
                    band UV absorbance (around 253.7 nm) are  confirmed interferences.
                    The concentration levels for interferants are difficult to define.
                    This suggests that quality control procedures (Sect. 10) must be
                    strictly followed.

               4.2  Volatile materials which absorb at 253.7  nm will cause a positive
                    interference.  In order to remove any interfering volatile
                    materials, the dead air space in the BOO  bottle should be purged
                    before the addition of stannous  chloride  solution.

               4.3  Interferences associated with the tissue  matrix are corrected for in
                    calibration procedure  (Sect. 9).
                                               283

-------
07/21/2003 ]0:28 FAX                                                               i]006/011
         5.   SAFETY

              5.1  The toxicity and cardnogenicity of each reagent used in this method
                   has not been fully established.  Each chemical should be regarded as
                   a potential health hazard and exposure to these compounds should be
                   minimized by good laboratory practices .   Normal  accepted
                   laboratory safety practices should be followed during reagent
                   preparation and instrument operation.  Always wear safety glasses or
                   full-face shield for eye protection when working with these
                   reagents.  Each laboratory is responsible for maintaining a current
                   safety plan, a current awareness file of OSHA regulations regarding
                   the safe handling of the chemicals specified in this method  '   .

              5.2  Mercury compounds are highly toxic if swallowed, inhaled, or
                   absorbed through the skin.  Analyses should be conducted in a
                   laboratory exhaust hood. The analyst should use chemical resistant
                   gloves when handling concentrated mercury standards.

              5.3  All personnel handling tissue samples should beware of biological
                   hazards associated with tissue samples.  Bivalve mollusk may
                   concentrate toxins and pathogenic organisms.  Tissue dissection
                   should be conducted in a bio-hazard hood and personnel should wear
                   surgical mask and gloves.

         6.   APPARATUS AND EQUIPMENT

              6.1  ABSORPTION CELL - Standard spectrophotometer cells 10-cm long,
                   having quartz windows may be used.  Suitable cells may be
                   constructed from plexiglass tubing, 1-in. O.p. by 4-1/2-in. long.
                   The ends are ground perpendicular to the longitudinal axis and
                   quartz windows (1-in- diameter by 1/16-in. thickness) are cemented
                   in place.  Gas inlet and outlet ports (also of plexiglass but 1/4-
                   in. O.D.) are attached approximately 1/2-in. from each end.  The
                   cell is strapped to a burner for support and aligned in the light
                   beam to give the maximum transmittance.

              6.2  AERATION TUBING - Inert mercury-free tubing is used for passage of
                   mercury vapor from the sample bottle to the absorption cell.   In
                   some systems, mercury vapor is recycled.   Straight glass tubing
                   terminating in a coarse porous glass aspirator is used for purging
                   mercury released from the tissue sample in the BOO bottle.

              6.3  AIR PUMP - Any pump (pressure or vacuum system) capable of passing
                   air at 1 L/min is used.  Regulated compressed air can be used in an
                   open one-pass system.

              6.4  ATOMIC ABSORPTION SPECTROPHOTOMETER - Any atomic absorption unit
                   having an open sample presentation area in which to mount the
                   absorption cell is suitable.  Instrument settings recommended by the
                   particular manufacturer should be followed.  Instruments designed
                   specifically for mercury measurement using the cold vapor technique


                                               284

-------
                                                                                  iioos
07/21/2003 10:24 FAX
                    are  commercially available and  may be substituted for the atomic
                    absorption  spectrophotometer.

               6.5   BIOCHEMICAL OXYGEN DEMAND (BOD)  BOTTLE -  See  Sect.  3.1.

               6.6   DRYING  TUBE -  Tube (6-in.  x 3/4-in.  OD) containing 20 g  of magnesium
                    perchlorate.   The filled  tube is Inserted (in-line) between the BOD
                    bottle  and  the absorption tube.   In  place of  the  magnesium
                    perchlorate drying tube,  a small  reading  lamp is  positioned to
                    radiate heat (about 10°C  above  ambient) on the absorption cell.
                    This avoids water condensation  in the cell.

               6.7   FLOWMETER - Capable of measuring an  air flow  of 1 L/min.

               6.8   MERCURY HOLLOW CATHODE LAMP - Single element  hollow cathode lamp or
                    electrodeless  discharge lamp and associated power supply.

               6.9   RECORDER -  Any raulti-range variable  speed recorder that  is
                    compatible  with the UV detection system is suitable.

               6.10  WATER BATH  - The water bath should have a covered top and capacity
                    to sustain  a water depth  of 2-1n.  to 3-in.  at 95*C ± 1°C.   The
                    dimensions  of  the water bath should  be large  enough to accommodate
                    BOD  bottles containing CAL,  LFB,  LFM,  LRB,  QCS and tissue samples
                    with the lid on.

               REAGENTS  AND CONSUMABLE MATERIALS

               7.1   Reagents may contain  elemental  impurities which bias  analytical
                    results.  All  reagents should be assayed  by the chemical
                    manufacturer for mercury  and meet ACS specifications.

                    7.1.1     Hydroxylamine Hydrochloride (NHjOH-HCl), (CASRN 5470-11-1)
                             may  be  used in  place  of hydroxylamine sulfate  in Sect.
                             7.6.   The assayed mercury  level  of  either compound is  not
                             to exceed 0.05  ppm.

                    7.1.2     Hydroxylamine Sulfate  [(NH2OH)2'H2SOJ (CASRN 10039-54-0);
                             assayed mercury level  1s not to exceed  1  ppb.

                    7.1.3     Mercuric Chloride (HgCl2),  (CASRN 7487-94-7).

                    7.1.4     Nitric  Acid (HN03), concentrated (sp.gr. 1.41),  (CASRN
                             7697-37-2);  assayed mercury level is  not  to  exceed 1  ppb.

                    7.1.5     Potassium Permanganate  (KMn04),  (CASRN 7722-64-7); assayed
                             mercury level 1s  not  to  exceed  0.05 ppm.

                    7.1.6     Potassium Persulfate  (K2S2Oa), (CASRN 7727-21-1); assayed
                             mercury level is  not to  exceed  0.05 ppm.

                    7.1.7     Reagent Water, ASTH type II.*

                                               285

-------
07/21/2003  10:28 F\X                                                               0005/011
                    7.1.8     Sodium Chloride (NaCl), (CASRN 7647-14-5); assayed mercury
                              level 1s not to exceed 0.05 ppm.

                    7.1.9     Stannous Chloride (SnCl2'2H20), (CASRN 10025-69-1);
                              assayed mercury level is not to exceed 0.05 ppra.

                    7.1.10    Stannous Sulfate (SnSOJ,  (CASRN 7488-55-3);  assayed
                              mercury level  is not to exceed 0.05 ppm,

                    7.1.11    Sulfurlc Acid (H2S04),  concentrated  (sp.gr. 1.84),  (CASRN
                              7664-93-9);  assayed mercury level is not to exceed 1 ppb.

               7.2  MERCURY CALIBRATION STANDARD - To each volumetric flask used for
                    serial dilutions, acidify with (0.1 to 0.256 by volume)  HNOj
                    (Sect. 7.1.4).   Using  mercury stock standard (Sect. 7.3), make
                    serial dilutions to obtain a concentration of 0.1 /*g Hg/mL.  This
                    standard should be prepared just before analyses.

               7.3  MERCURY STOCK STANDARD - Dissolve in a 100-tnL volumetric flask
                    0.1354 g HgCl,  (Sect.  7.1.3)  with 75 ml of reagent water
                    (Sect. 7.1.7).   Add 10 ml of cone. HN03 (Sect.  7.1.4) and dilute to
                    mark.   Concentration is 1.0 mg Hg/mL.

               7.4  POTASSIUM PERMANGANATE SOLUTION - Dissolve 5 g of KMn04
                    (Sect. 7.1.5) in 100 ml of reagent water (Sect. 7,1.7).

               7.5  POTASSIUM PERSULFATE SOLUTION - Dissolve 5 g of K2S208 (Sect. 7.1.6)
                    in 100 mL of reagent water (Sect. 7,1.7).

               7.6  SODIUM CHLORIDE-HYDROXYLAMINE SULFATE SOLUTION - Dissolve 12 g of
                    NaCl  (Sect.  7.1.8) and 12 g of (NH2OH)2'H2SO,  (Sect.  7.1.2)  or 12 g
                    of NHjOH'HCl  (Sect. 7.1.1) dilute with  reagent water (Sect.  7.1.7)
                    to 100 mL.

               7.7  STANNOUS CHLORIDE SOLUTION - Add 25 g SnCl2'2H20 (Sect.  7.1.9) or
                    25 g  of SnS04 to 250 raL  of 0.5 N H2S04 (Sect. 7.8).  This mixture is
                    a suspension and should be stirred continuously during  use.

               7.8  SULFURIC ACID,  0.5 N - Slowly add 14.0 mL  of cone. H2SO.
                    (Sect. 7.1.10)  dilute  to 1 L with reagent  water (Sect.  7.1.7).

          8.    SAMPLE COLLECTION. PRESERVATION AND STORAGE

               8.1  Because of the  extreme sensitivity of the  analytical procedure and
                    the presence of mercury in a laboratory environment, care must be
                    taken to avoid  extraneous contamination.  Sampling devices, sample
                    containers  and  plastic items should be determined to be free of
                    mercury; the sample should not be exposed  to any condition in the.
                    laboratory that may result In contact or airborne mercury
                    contamination.
                                               286

-------
07/21/2003  10:24 FAX
             8.2  The  tissue  sample  should  be  preserved and dissected  In  accordance
                  with Method 200.3,  "Sample Preparation Procedure  for Spectrochemical
                  Determination  of Total Recoverable  Elements  in  Biological  Tissues",
                  only Sect.  8.  Tissue Dissection,  is used in  this  method.

             8.3  Weigh 0.2-  to  0.3-g portions of each sample  and place in the bottom
                  of a dry BOD bottle.  Care must be  taken that none of the  sample
                  adheres to  the side of the bottle.  Immediately cap  and cover the
                  top  of the  BOO bottle with aluminum foil.

        9«   CALIBRATION AND  STANDARDIZATION

             9.1  The  calibration curve is  prepared from values determined for
                  portions of fortified tissue treated in the  manner used for the
                  tissue samples being analyzed.  For preparation of the  calibration
                  standards,  blend a  portion of tissue in a Waring  blender.

             9.2  Transfer accurately weighed portions to each of five dry BOD
                  bottles.  Each sample should weigh  about 0.2 g.   Add 4  ml  of cone.
                  H2S04 and  1  ml  of cone.  HN03  to each bottle and  place in a  water
                  bath maintained at  58°C until the tissue is  completely  dissolved (30
                  to 60 minutes).

             9.3  Cool and transfer 0.5, 2.0, 5.0 and 10.0 mL  aliquots of the CAL
                  (Sect. 7.2)  solution containing 0.5 to 1.0 fig of  Hg  to  the BOD
                  bottles containing  tissue.  Cool to 4°C in an ice bath  and
                  cautiously  add 15 ml of potassium permanganate  solution (Sect. 7.4)
                  and  8 ml of potassium persulfate (Sect. 7.5).   Allow to stand
                  overnight at room temperature under oxidizing conditions.

             9.4  Construct a standard curve by plotting peak  height or maximum
                  response of the standard  (obtained  in Sect,  11.7) versus micrograms
                  of mercury  contained in the bottles. The standard curve should
                  comply with  Sect. 10.2.3.  Calibration using computer or calculator
                  based regression curve fitting techniques on concentration/response
                  data is acceptable.

        10.  QUALITY CONTROL

             10.1 Each laboratory using this method is required to  operate a formal
                  quality control (QC) program.  The minimum requirements of this
                  program consist of  an initial demonstration of  laboratory capability
                  by analyses  of laboratory reagent blanks, fortified blanks and
                  samples used for continuing check on method performance.  Standard
                  Reference Materials (SRMs) •  6 are available  and should  be  used to
                  validate laboratory performance.  Commercially  available tissue
                  reference materials are acceptable for routine  laboratory use.  The
                  laboratory  is  required to maintain  performance records that define
                  the  quality of data generated.
                                              287

-------
07/21/2003 10:27 FAX                                                               i]004/011
              10,2  INITIAL DEMONSTRATION  OF  PERFORMANCE

                   10.2.1    The  Initial  demonstration of performance  Is used  to
                            characterize Instrument performance  (MOLs  and  linear
                            calibration  ranges)  for analyses conducted by  this method.

                   10.2.2    A mercury HDL should be established  using  LFH  at  a
                            concentration of two to five times the estimated  detection
                            limit7.  To  determine MDL values, take seven replicate
                            aliquots of  the LFH  and process through the entire
                            analytical method.   Perform all calculations defined 1n
                            the  method and  report the concentration values in the
                            appropriate  units.   Calculate the MDL as  follows:

                            MDL  -  (t) x  (S)

                            where,  t ->   Student's t value for a  99X confidence level
                                         and a standard deviation estimate  with n-1
                                         degrees  of freedom [t =  3.14  for seven
                                         replicates].

                                    S =   standard deviation of the replicate analyses.

                            A MDL  should be determined every six months or whenever a
                            significant  change in background or  instrument response 1s
                            expected (e.g., detector change).

                   10.2.3    Linear calibration ranges - The upper limit of the linear
                            calibration  range should be established for mercury by
                            determining  the signal responses from a minimum of three
                            different concentration standards, one of  which is close
                            to the upper limit of the linear range. Linear calibration
                            ranges should be determined every six months or whenever a
                            significant  change in Instrument response  is observed.

              10.3  ASSESSING LABORATORY PERFORMANCE - REAGENT AND FORTIFIED BLANKS

                   10.3.1    The  laboratory  must  analyze at least one  LRB (Sect. 3.7)
                            with each set of samples.  LRB data  are used to assess
                            contamination from the laboratory environment  and to
                            characterize spectral background from the  reagents used in
                            sample processing.   If an mercury value in a LRB  exceeds
                            its  determined  MDL,  then laboratory  or reagent
                            contamination is suspect.  Any determined  source  of
                            contamination should be corrected and the  samples
                            reanalyzed.

                   10.3.2    The  laboratory  must  analyze at least one LF6 (Sect. 3.5)
                            with each batch of samples.  Calculate accuracy as percent
                            recovery (Sect. 10.4.2).  If the recovery  of mercury falls
                            outside control limits (Sect. 10.3.3), the method is
                                              288

-------
                                                                                  iioio
07/21/2003 10:25 FAX
                             judged out of control.   The source of the problem should
                             be Identified and resolved before continuing analyses.

                   10.3.3    Until sufficient data (usually a  minimum of 20 to 30
                             analyses)  become available, each  laboratory should assess
                             its performance against recovery  limits of 85-115%.   When
                             sufficient internal  performance data become available,
                             develop control limits  from the percent mean recovery (x)
                             and the standard deviation (S) of the mean recovery.
                             These data are used  to  establish  upper and lower control
                             limits as  follows:

                                UPPER CONTROL LIMIT  - x -f 3S
                                LOWER CONTROL LIMIT  - x - 3S

                             After each five to ten  new recovery measurements,  new
                             control limits should be calculated using only the most
                             recent 20  to 30 data points.

              10.4  ASSESSING ANALYTE RECOVERY - LABORATORY FORTIFIED SAMPLE MATRIX

                   10.4.1    The laboratory must  add a known amount of mercury to a
                             minimum of 10% of samples or one  sample per sample set,
                             whichever  is greater.   Select a tissue sample that is
                             representative of the type of tissue being analyzed  and
                             has a low  mercury background.   It is recommended that this
                             sample be  analyzed prior to fortification.   The
                             fortification should be 20% to 50% higher than the
                             analyzed value.  Over time,  samples from all  routine
                             sample sources should be fortified.

                   10.4.2    Calculate  the percent recovery, corrected for background
                             concentrations measured in the unfortified sample, and
                             compare these values to the control  limits established in
                             Sect.  10.3.3 for the analyses  of  LFBs.   A recovery
                             calculation is not required if the concentration of  the
                             analyte added is less than 10% of the sample background
                             concentration.   Percent recovery  may be calculated in
                             units appropriate to the matrix,  using the following
                             equation:

                                 C  - C
                             R  =  *	  x 100
                            where, R   = percent recovery
                                   Cs  = fortified sample concentration
                                   C   = sample background concentration
                                   s   » concentration equivalent of
                                        fortifier added to tissue sample.
                                              289

-------
07/21/2003  10:27 FAX                                                               0003/011
                  10.4.3    If mercury recovery falls outside the designated range,
                            and the laboratory performance Is shown to be In control
                            (Sect.  10.3),  the recovery problem encountered with the
                            fortified tissue sample 1s judged to be matrix related,
                            not system related.  The result for mercury 1n the
                            unfortified sample must be labelled to Inform the data
                            user that the results are suspect due to matrix effects.

        11.   PROCEDURE

             11.1  Add 4 mL of cone. H2SO,  (Sect.  7.1.10} and  1 ml of cone, HNO,
                  (Sect. 7.1.4) to each Dottle and place In a water bath maintained at
                  58°C until  the tissue is completely dissolved (30 to 60 min).

             11.2  Cool to 4°C in an ice bath and cautiously add 5 ml of potassium
                  permanganate solution (Sect. 7.4) in 1 mL increments.  Add an
                  additional  10 ml or more of permanganate, as necessary to maintain
                  oxidizing conditions.  Add 8 ml of potassium persulfate solution
                  (Sect. 7.5).  Allow to stand overnight at room temperature.

                  As an alternative to the overnight digestion, tissue solubilization
                  may be carried out In a water bath at 80°C for 30 min.  The sample
                  is cooled and IB mL of potassium permanganate solution (Sect. 7.4)
                  added cautiously followed by 8 ml of potassium persulfate solution
                  (Sect, 7.5).  At this point, the sample is returned to the water
                  bath and digested for an additional 90 min at 30°C.   Calibration
                  standards are treated in the same manner.

             11.3  Turn on the spectrophotometer and circulating pump.   Adjust the pump
                  rate to 1 L/min.   Allow the spectrophotometer and pump to stabilize.

             11.4  Cool the 300 bottles to room temperature and dilute in the following
                  manner:

                  11.4.1    To each BOO bottle containing the CAL,  LFB and LRB, add 50
                            mL of reagent water (Sect. 7.1.7).

                  11.4.2    To each BOD bottle containing a tissue sample, QCS or LFM,
                            add 55 mL of reagent water (Sect. 7.1.7).

             11.5  To each BOO bottle, add 6 mL of sodium chloride-hydroxylamine
                  sulfate solution (Sect.  7.6) to reduce the excess permanganate.

             11.6  Treating each bottle individually:

                  11.6.1    Placing the aspirator Inside the BOD bottle and above the
                            liquid, purge the head space (20 to 30 sec) to remove
                            possible gaseous interferents.

                  11.6.2    Add 5 mL of stannous chloride solution (Sect. 7.7) and
                            Immediately attach the bottle to the aeration apparatus.


                                              290

-------
07/21/2003 10:25 FAX
                  11.6.3    The absorbance,  as exhibited either on the spectro-
                            photometer or the recorder,  will  increase  and  reach
                            maximum within 30 sec.   As  soon  as  the recorder pen levels
                            off,  approximately 1  min, open the  bypass  value (or
                            optionally remove aspirator from the BOD bottle if it Is
                            vented under the hood)  and  continue the aeration until  the
                            absorbance returns to its minimum value.

             11.7 Close the bypass value, remove  the aspirator  from the BOD bottle  and
                  continue the aeration.  Repeat  step  (Sect.  11.6) until all  BOD
                  bottles have been aerated and recorded.

        12.   CALCULATIONS

             12.1 Measure the peak height of the  unknown from the chart and read the
                  mercury value from the standard curve.

             12.2 Calculate the mercury concentration 1n the sample by the formula:


                         Wtf Hg/g =      \ig fig  in the aliquot
                                     wt.  of  the aliquot  in grams

             12.3 Report mercury concentrations as  follows:   Below 0.1  M9/9.  <
                  0.1 /jg/g; between 0.1 and 1 M9/9» to  the  nearest 0.01 ^g; between 1
                  and 10 Atg/g, to nearest 0.1 /itg; above 10 /^g/g,  to nearest M9-
        13.   PRECISION AND ACCURACY

             13.1  The standard deviation for mercury  in  fish  tissue  samples  are
                  reported as  0.19 ± 0.02 fig Hg/g  , 0.74 ± 0.05  ng Hg/g  and  0.74 ±
                  0.05 ng Hg/g with recoveries  for LFM being  112%, 93%,  and  86%,
                  respectively.   These tissue samples  were fortified with methyl
                  mercuric chloride.

        14.   REFERENCES

             1.    "Safety in Academic Chemistry Laboratories", American  Chemical
                  Society Publication, Committee on Chemical  Safety, 3rd Edition,
                  1979,

             2.    "OSHA Safety and Health Standards, General  Industry",  (29CFR 1910),
                  Occupational  Safety and Health Administration, OSHA 2206,  revised
                  January,  1976,

             3.    "Proposed OSHA  Safety and  Health Standards, Laboratories",
                  Occupational  Safety and Health Administration, Federal Register,
                  July 24,  1986.

             4.    "Specification  for Reagent Water," Annual Book of  ASTM Standards,
                  D1193,  Vol.  11.01, 1990.


                                             291

-------
07/21/2003 10:27 FAX                                                               i]002/0il
             5.    National  Institute of Standards and Technology,  Office of Standards
                  Reference Materials,  Gaithersburg,  MD Z0899:   Aquatic Plant -
                  Lagarosiphon  major (CRM 8030),  Aquatic Plant  - Platihypnidium
                  riparioldes (CRM 8031), Oyster  Tissue (SRM 1566a),  Albacore Tuna
                  (RM 50).

             6.    National  Research Council  of Canada,  Marine Analytical Chemistry
                  Standards Program,  Division of  Chemistry,  Montreal  Road,  Ottawa,
                  Ontario K1A OR9, Canada:  Dogfish Liver (DOLT-1),  Dogfish Muscle
                  (DORM-1), Non Defatted Lobster  Kepatopancreas (LUTS-1), Lobster
                  Hepatopancreas (TORT-1).

             7.    Code of Federal  Regulations 40, Ch. 1, Pt. 136 Appendix B.
                                             292

-------
07/21/2003 10:25 FAX
                                                                              21012
                  0 * BUBBLER
ABSORPTION
    CELL
           SAMPLE  SOLUTION
           IN BOO  BOTTLE
SCRU88ER
CONTAINING
A MERCURY
A8SORBNO
MEDIA
                      1.  Apparatus f«r Fl «*!•$$ torcury
      Because  of the toxic nature of mercury vapor,  Inhalation must be avoided.
      Therefore, a bypass has been included in the system to either vent the mercury
      vapor into a exhaust hood or pass the vapor through some absorbing media,  such
      as:    a) equal volumes of 0.1 N KHnO, and 10X H2SO,
            b) 0.25% iodine In a 3X KI solution.
      A specially treated charcoal  that will absorb mercury vapor is also available
      from Barnebey and Cheney,  P.O. Box 2526, Columbus, OH  43216, Catalog No.  580-
      13 or 580-22.
                                         293
                                                               •U.S. Sovemnuin Printing OTBcK 1991 — S4S-187/40651

-------