United States
                   Environmental Protection
                   Agency
                   Environmental Monitoring and
                   Support Laboratory
                   Cincinnati OH 45268
                   Research and Development
                                       EPA-600/4-82-055 Dec. 1982
&EPA
Test Methods
                   Technical Additions to
                   Methods for Chemical
                   Analysis of Water and
                   Wastes

-------
                              Notice
This  document  has  been reviewed in  accordance  with  U.S.  Environmental
Protection Agency policy and approved for publication. Mention of trade names or
commercial products does not constitute endorsement or recommendation for use.

-------
                              Foreword
Environmental measurements are required to determine the quality of
ambient waters and the character of waste effluents. The Environmental
Monitoring and Support Laboratory - Cincinnati, conducts research to:
• Develop and evaluate methods to measure the presence  and concentration
of physical, chemical, and radiological  pollutants in water, wastewater, bottom
sediments, and solid waste
• Investigate methods  for the concentration, recovery, and  identification of
viruses, bacteria and other microbiological organisms  in water; and, to
determine the responses of aquatic organisms to water quality.
• Develop and operate an Agency-wide quality assurance program to assure
standardization and quality control of systems for monitoring water and
wastewater.
• Develop and operate a computerized system for instrument automation
leading to improved data collection, analysis, and quality control.
The preservation and holding times table  and the six methods herein  have been
added to Methods for Chemical Analysis of Water and Wastes, EPA 600/4-79-
020, as of the second printing, in an effort to provide updated analytical
information to laboratories complying with the Safe Drinking Water Act, the
National Pollutant Discharge Elimination System, Section 304(h) of the Clean
Water  Act, and the Ambient Monitoring Requirements of Sections 106 and 208
of Public Law 92-500.
                                     Robert L. Booth, Acting Director
                                     Environmental Monitoring and Support
                                     Laboratory - Cincinnati

-------
Table 1 . Recommendation for Sampling and Preservation of Samples
According to Measurement*
Vol.
Req Holding
Measurement (mL) Container2 Preservative3" Time5
100 Physical Properties
Color
Conductance
Hardness
Odor
pH

Residue
Filterable
Non-
Filterable
Total
Volatile
Settleable
Matter
Temperature

Turbidity
200 Metals
Dissolved

Suspended
Total
Chromium*6
Mercury
Dissolved

Total
50
100
WO
200
25


WO

WO
WO
WO
WOO

WOO

WO

200

200
WO
200

WO

WO
P,G
P.G
P.G
G only
P.G


P.G

P.G
P.G
P.G
P.G

P.G

P.G

P.G


P.G
P.G

P.G

P.G
Cool, 4°C
Cool. 4°C
HN03 to pH<2
Cool, 4°C
None Req.


Cool, 4°C

Cool, 4°C
Cool, 4°C
Cool. 4°C
Cool, 4°C

None Req.

Cool, 4°C

Filter on site
HNOy, to pH <2
Filter on site
HN03 to pH <2
Cool, 4°C

Filter
HN03 to pH <2
HNO3 to pH <2
48 Hrs.
28 Days
6 Mos.
24 Hrs.
Analyze
Immediately

7 Days

7 Days
7 Days
7 Days
48 Hrs.

Analyze
Immediately
48 Hrs.

6 Mos.8

6 Mos.
6 Mos.
24 Hrs.

28 Days

28 Days
300 Inorganics, Non-Metallics
Acidity
Alkalinity
Bromide
Chloride
Chlorine

Cyanides


Fluoride
Iodide
Nitrogen
Ammonia

Kjeldahl, Total

Nitrate plus
Nitrite

Nitrate*
Nitrite
Dissolved Oxygen
Probe

Winkler


Phosphorus
Ortho-
phosphate,
Dissolved
Hydrolyzable

Total
WO
WO
WO
50
200

500


300
WO

400

500


WO

WO
50

300

300




50

50

50
P.G
P.G
P.G
P.G
P.G

P.G


P.G
P.G

P.G

P.G


P.G

P.G
P.G

G bottle and top

G bottle and top




P.G

P.G

P.G
Cool, 4°C
Cool, 4°C
None Req.
None Req.
None Req.

Cool. 4°C
NaOH to pH>J2
0.6g ascorbic acid6
None Req.
Cool, 4°C

Cool, 4°C
/H2SO4 to pH <2
Cool, 4°C
H2SO, to pH <2

Cool, 4°C
#2504 to pH <2
Cool, 4°C
Cool, 4°C

None Req.

Fix on site
and store
in dark


Filter on site
Cool, 4°C
Cool, 4°C
HzSO* to pH <2
Cool, 4°C
14 Days
14 Days
28 Days
28 Days
Analyze
Immediately
14 Days7


28 Days
24 Hrs.

28 Days

28 Days


28 Days

48 Hrs.
48 Hrs.

Analyze
Immediately
8 Hours




48 Hrs.

28 Days

28 Days

-------

Total,
Dissolved

Silica
Suit ate
Sulfide



Sulfite
400 Organics
BOD
COD

Oil & Grease


Organic carbon

Pheno/ics

MB AS
NTA

50


50
50
500



50

WOO
50

WOO


25

500

250
50

P.G


P only
P.G
P.G



P.G

P.G
P.G

G only


P.G

G only

P.G
P.G
H2SO* to pH <2
Filter on site
Cool, 4°C
HzSO* to pH <2
Cool, 4°C
Cool, 4°C
Cool, 4°C
add 2 mL zinc
acetate plus NaOH
to pH >9
None Req.

Cool, 4°C
Cool, 4°C
H2SO* tp pH <2
Cool. 4°C
H2SO*
topH <2
Cool, 4°C HCI
or H2S04 to pH <2
Cool, 4°C
H2S04 to pH <2
Cool, 4°C
Cool, 4°C

24 Hrs.


28 Days
28 Days
7 Days



Analyze
Immediately
48 Hrs.
28 Days

28 Days


28 Days

28 Days

48 Hrs.
24 Hrs.
1 More specific instructions for preservation and sampling are found with each
procedure as detailed in this manual. A general discussion on sampling water and
industrial wastewater may be found in ASTM, Part 31, p.  72-82 (1976) Method D-
3370.
2 Plastic (P) or Glass (G). For metals, polyethylene with a polypropylene cap (no liner)
is preferred.
3 Sample preservation should be performed immediately upon sample collection.
For composite samples each aliquot should be preserved at the time of collection.
When use of an automated sampler makes it impossible to preserve each aliquot,
then samples may be preserved by maintaining at 4°C  until compositing and
sample splitting are completed.
*When any sample is to be shipped by common carrier or sent through the United
States Mails, it must comply with the Department of Transportation Hazardous
Materials Regulations (49 CFR Part 172). The person offering such material for
transportation is responsible for ensuring such compliance. For the preservation
requirements of Table  1, the Office  of Hazardous Materials, Materials
Transportation Bureau, Department of Transportation has determined that the
Hazardous Materials Regulations do not  apply  to the following materials:
Hydrochloric acid (HCIj in  water solutions at concentrations of 0.04% by weight or
less (pH about  1.96  or greater); Nitric acid (HNOs) in  water solutions  at
concentrations of 0.15% by weight or less fpH about 1.62 or greater); Suit uric acid
(HzSOt) in water solutions at concentrations of 0.35% by weight or less (pH about
1.15 or greater); and Sodium hydroxide (NaOH) in water solutions at concentrations
of 0.080% by weight or less (pH about 12.30 or less).
5 Samples should be analyzed as soon as possible after collection.  The times listed
are the  maximum times that samples may be  held before analysis  and still
considered valid. Samples may be held for longer periods only if the permittee, or
monitoring laboratory, has data on file to show that the specific types of samples
under study are stable for the longer time, and has received a variance from the
Regional Administrator. Some samples may not be  stable  for the maximum time
period given in the table. A permittee, or monitoring laboratory, is obligated to hold
the sample for a shorter time if know/edge exists  to show  this is necessary to
maintain sample stability.
6 Should only be used in the presence of residual chlorine.
7 Maximum ho/ding  time is 24  hours  when sulfide is present.  Optionally, all
samples may be tested with lead acetate paper before the pH adjustment in order to
determine if sulfide is present. If sulfide is present, it can be removed by the addition
of cadmium nitrate powder until a negative spot  test is obtained. The sample is
filtered and then NaOH is added to pH 12.
a Samples should be filtered immediately on-site adding preservative for dissolved
metals.
9For samples from non-chlorinated drinking water sup/ies  cone. H2SO* should be
added to lower sample pH to less than 2. The sample should be analyzed before 14
days.

-------
              Environmental Protection  Agency
         Regional Quality Assurance Coordinators
Region 1
Warren H. Oldaker
Central Regional Laboratory
Environmental Services Division
U.S. Environmental Protection Agency
60 Westview Street
Lexington, MA 02173
(617-861-6700)
FTS 8-617-861-6700
Region 2
Gerard F. McKenna
Research and QualityAssurance Branch
Environmental Services Division
U.S. Environmental Protection Agency
Edison, NJ 08837
(201-321-6645)
FTS 340-6645
Region 3
Charles Jones,  Jr.
(3SA60)
Water Quality Monitoring Branch
Environmental Services Division
U.S. Environmental Protection Agency
6th & Walnut Streets, Curtis Bldg.
Philadelphia, PA 19106
(215-597-9162)
FTS 597-9162
Region 4
Wade Knight
Laboratory Services Branch
Environmental Services Division
US. Environmental Protection Agency
College Station Road
Athens, GA 30613
(404-546-3165)
FTS 250-3165
Region 5
David Payne
Quality Assurance Office
Environmental Services Division
U.S. Environmental Protection Agency
536 South Clark Street
Chicago, IL 60605
(312-353-7712)
FTS 353-7712
Region 6
Eloy R. Lozano
Environmental Services Division
U.S. Environmental Protection Agency
1201 Elm St.,
First Int'l Bldg.
Dallas, TX 75270
(214-767-2697)
FTS 729-2697
Region 7
Charles P. Hensley
Environmental Services Division
U.S. Environmental Protection Agency
25 Funston  Road
Kansas City, KS 66115
(816-374-4285)
FTS 758-4285
Region 8
Dr. Juanita  Hillman
Environmental Services Division
U.S. Environmental Protection Agency
Lincoln Tower Bldg , Suite 900
1860 Lincoln St.
Denver, CO  80295
(303-327-4935)
FTS 327-4935
Region 9
Dr. Ho Young
Office of Quality Assurance  and Monitoring Systems
U.S. Environmental Protection Agency
215 Fremont St.,
San Francisco, CA 94105
(415-556-2647)
FTS 556-2647
Region 10
Barry Townes
Environmental Services Division
U.S. Environmental Protection Agency
1200 Sixth Avenue
Seattle, WA 98101
(206-442-1675)
FTS 399-1675

-------
                                  United States
                                  Environmental Protection
                                  Agency
                                  Environmental Monitoring and
                                  Support Laboratory
                                  Cincinnati OH 45268
                                 Research and Development
SEPA
Test  Method
                                 pH,  Continuous  Monitoring
                                 (Electrometric)—Method  150.2
                                 1.  Scope and Application

                                 1.1 This method is applicable to the
                                 continuous pH measurement of
                                 drinking, surface, and saline waters,
                                 domestic and industrial waste waters.

                                 2.  Summary of Method

                                 2.1 The pH of a sample is determined
                                 electrometrically using a glass
                                 electrode with a reference electrode
                                 or a single combination electrode.

                                 3.  Sample Handling and
                                 Preservation

                                 3.1 The composition of the water or
                                 waste contacting the measuring
                                 electrode system must be
                                 representative of the total flow from
                                 the water body.

                                 4.  Interferences

                                 4.1 The glass electrode, in general,
                                 is not subject to solution interferences
                                 from color, turbidity, colloidal matter,
                                 oxidants, reductants or high salinity.

                                 4.2 Sodium error at pH levels
                                 greater than 10 can be reduced or
                                 eliminated by using a "low sodium
                                 error" electrode.

                                 4.3 Manually inspect the conditions
                                 of the electrodes every 30 days for
                                 coating by oily materials or buildup of
                                 lime. If oil and grease and/or scale
                                 buildup are not present, this time
                                 interval  may be extended.

                                 4.3.1  Coatings of oil, grease and
                                 very fine solids can impair electrode
                                 response. These can usually be
                                 removed by gentle wiping and
                                 detergent washing. The use of flow-
                                 through electrode housings which
                                 provide higher flow velocity helps to
                                 prevent the coating action.

                                 4.3.2  Heavy particulate matter such
                                 as lime accumulation can be removed
                                 by careful scrubbing or immersion in
                                 dilute (1+9) hydrochloric acid.
                                 Continuous monitoring under these
                                 conditions benefits from ultrasonic or
                                 other in-line continuous cleaning
                                 methods.

                                 4.4  Temperature effects on the
                                 electrometric measurement of pH
                                 arise from two sources. The first is
                                 caused by the change in electrode
                                 output at various temperatures. This
                                 interference can be controlled with
                                 instruments having temperature
                                 compensation or by calibrating the
                                 electrode-instrument system  at the
                                 temperature of the samples.  For best
                                 results, meters having automatic
                                 temperature compensation should be
                                 calibrated with solutions within 5°C of
                                 the temperature  of the stream to be
                                 measured. The second source is the
                                 change of pH inherent in the sample
                                 at various temperatures. This error is
                                 sample dependent and cannot be
                                 controlled, it should therefore be
                                 noted by reporting both the pH and
                                 temperature at the time of analysis.


                                 5.  Apparatus

                                 5.1   pH Monitor - A wide variety of
                                 instruments are commercially
                                 available with various specifications
                                 and optional equipment. For
                                 unattended  use, the monitor  should
                                 be equipped with automatic or fixed
                                 J 50.2-1
                                                         Dec. 1982

-------
temperature compensation and with a
recorder or alarm function.

5.2   Glass electrode - with shielded
cable between electrode and monitor
unless preamplification is used.

5.3   Reference electrode - a
reference electrode with a constant
potential and with either a visible
electrolyte or viscous gel fill.
NOTE 1:  Combination electrodes  in-
corporating both measuring and refer-
ence functions  are convenient to use
and are available with solid, gel-type fil-
ling   materials  that  require  minimal
maintenance.

5.4   Temperature sensor - for
automatic compensator covering
general ambient temperature range.

5.5   Electrode mounting - to hold
electrodes; may be flow through (for
small flows), pipe mounted or
immersion.

6.    Reagents

6.1   Primary  standard buffer salts are
available from the National Bureau  of
Standards and should be used in
situations where extreme accuracy  is
required.

6.1.1  Preparation of reference
solutions from the.se salts require
some special precautions and
handling1 such as low conductivity
dilution water, drying ovens, and
carbon dioxide free purge gas. These
solutions should be replaced at least
once each month.

6.2   Secondary buffers  may be
prepared from NBS salts or purchased
as a solution from commercial
vendors. Use of these commercially
available solutions, which have been
validated by comparison to NBS
standards, is recommended for
routine operation. These buffers may
be retained for at least six months if
kept stoppered.
7.   Calibration

7.1   Immersion type electrodes -
easily removed from mounting.

7. /. 7  The electrode should be
calibrated at a minimum of two points
that bracket the expected pH of the
water/waste and are approximately
three pH units or more apart.
'National Bureau of Standards Special Publication
260.
 7.1.2   Repeat calibration
 adjustments on successive portions of
 the two buffer solutions until readings
 are within ±0.05 pH units of the buffer
 value. If calibration problems occur,
 see 4.3.

 7.7.3   Because of the wide variety of
 instruments available, no detailed
 operating instructions are provided.
 Instead, the analyst should refer to
 the particular  manufacturer's
 instructions.

 7. 7.4   Calibration against two buffers
 should be carried out at least daily.  If
 the pH of the fluid being  measured
 fluctuates considerably, the  calibration
 should be carried out more often.
 Calibration frequencies may be
 relaxed if historical data supports a
 longer period between  calibration.

 7.2  Immersion type electrodes -
 not easily removed from  mounting.

 7.2.1   Collect a grab sample of the
 flowing material from a point as close
 to the electrode as possible. Measure
 the pH of this  grab sample as quickly
 as possible with a laboratory - type pH
 meter. Adjust  the calibration control
 of the continuous monitor to the
 reading obtained

 7.2.2  The temperature and condition
of the grab sample must remain
constant until  its pH has been
measured by the laboratory pH meter
The temperature of the sample should
be measured and the temperature
compensator of the laboratory pH
meter adjusted.

7.2.3  The laboratory - type  pH meter
should be calibrated prior to  use
against two buffers as outlined in 7.1.

 7.2.4  The continuous pH monitoring
system should be initially calibrated
against two buffers as outlined in 7.1
before being placed  into service.
Recalibration (every 30 days) at two
points is recommended if at  all
possible to ensure the measuring
electrode is in  working  order. If this is
not possible, the use of electrode
testing features for a broken or
malfunctioning electrode  should be
considered when purchasing the
equipment.

 7.2.5  The indirect  calibration should
be carried out  at least once a day.  If
the pH  of the fluid being measured
fluctuates considerably, the calibration
should be carried out more often.
Calibration frequencies may  be
relaxed  if historical data support a
longer period between calibration
 7.2,6  If the electrode can be
 removed from the system, but with
 difficulty, it should be directly
 calibrated as in 7.1 at  least once a
 month.

 7.3  Flow-through type electrode -
 easily removed from its mounting.

 7.3.1  Calibrate using buffers as in
 7.1. The buffers to be used may be
 the process stream itself as one
 buffer, and as a second buffer after
 adjustment of pH by addition of an
 acid or base. This will provide the
 larger volumes necessary to calibrate
 this type electrode.

 7.3.2  Since the velocity of sample
 flow-through a flow through electrode
 can  produce an offset error in pH
 reading, the user must have data on
 hand to show that the  offset is known
 and compensation has been
 accomplished.

 7.4  Flow-through type electrode -
 not easily removed from its mounting.

 7.4. 1  Calibrate as  in  7.2.

 7.4.2  Quality control data must be
 on hand to show the user is aware  of
 possible sample flow velocity effects.

 8.    Procedure

 8.1   Calibrate the monitor and
 electrode system as  outlined in
 Section 7.

 8.2   Follow the manufacturer's
 recommendation for operation and
 installation of the system.

 8.3   In wastewaters, the electrode
 may require periodic cleaning. After
 manual cleaning, the electrode should
 be calibrated as in 7.1  or 7.2 before
 returning to service.

 8.4   The electrode must be placed so
 that the water or waste flowing past
 the electrode is representative of the
 system.

 9.   Calculations

 9.1   pH meters read directly in pH
 units. Reports pH to  the nearest  0.1
 unit and temperature to the nearest
10.   Precision and Accuracy

10.1   Because of the wide variability
of equipment and conditions and the
changeable character of the pH of
many process waters and wastes, the
precision of this method is probably
less than that of Method 1 50. 1 ;
however, a precision of 0.1 pH unit
                                      Dec. 1982
                                                                    150.2-2

-------
should be attainable in the range of
pH 6.0 to 80. Accuracy data for
continuous monitoring equipment
are not available at this time.
Bibliography
1.  Annual Book of ASTM
Standards, Part 31, "Water"
Standard 1293-78, Method D, p. 226
(1981).
                                     150.2-3                     Dec. 1982

-------
                               United States
                               Environmental Protection
                               Agency
                               Environmental Monitoring and
                               Support Laboratory
                               Cincinnati OH 45268
                               Research and Development
&EPA
Test  Method
                               Inductively Coupled  Plasma—
                              Atomic Emission  Spectrometric
                               Method for Trace  Element
                              Analysis  of Water  and
                              Wastes—Method  200.7
                               1.  Scope and Application

                               1.1 This method may be used for
                               the determination of dissolved,
                               suspended, or total elements in
                               drinking water, surface water,
                               domestic and industrial wastewaters.

                               1.2 Dissolved elements are
                               determined in filtered and acidified
                               samples. Appropriate steps must be
                               taken in all analyses to ensure that
                               potential interference are taken into
                               account. This is especially true when
                               dissolved solids exceed 1 500 mg/L.
                               (See 5.)

                               1.3 Total elements are determined
                               after appropriate digestion procedures
                               are performed. Since digestion
                               techniques increase the dissolved
                               solids content of the samples,
                               appropriate steps must be taken to
                               correct for potential interference
                               effects. (See 5.)
                               1.4 Table 1 lists elements for which
                               this method applies along with
                               recommended wavelengths and
                               typical estimated instrumental
                               detection limits using conventional
                               pneumatic nebulization. Actual
                               working detection limits are sample
                               dependent and as the sample matrix
                               varies, these concentrations may also
                               vary. In time, other elements may be
                               added as more information becomes
                               available and as required.

                               1.5  Because of the differences
                               between various makes and models of
                               satisfactory instruments, no detailed
                               instrumental  operating instructions
                               can be provided. Instead, the analyst
                               is referred to the instructions provided
                               by the manufacturer of the particular
                               instrument.

                               2.   Summary of Method

                               2.1  The method describes a
                               technique for the simultaneous or
                               sequential multielement
                               determination of trace elements in
                               solution. The basis of the method  is
                               the measurement of atomic emission
                               by an optical  spectroscopic technique.
                               Samples are  nebulized and the
                               aerosol that is produced is transported
                               to the plasma torch where excitation
                               occurs. Characteristic atomic-line
                               emission spectra are produced  by a
                               radio-frequency inductively coupled
                               plasma (ICP).  The spectra are
                               dispersed by a grating spectrometer
                               and the  intensities of the lines  are
                               monitored by photomultiplier tubes.
                               The photocurrents from the
                               photomultiplier tubes are processed
                               and controlled by a computer system.
                               A background correction technique is
                               required to compensate for variable
                               background contribution to the
                              Meta/s-1
                                                   Dec. 1982

-------
determination of trace elements.
Background must be measured
adjacent to analyte  lines on samples
during analysis. The position selected
for the background  intensity
measurement, on either or both sides
of the analytical line, will be
determined by the complexity of the
spectrum adjacent to the analyte line.
The position used must  be free of
spectral interference and reflect the
same change in background
intensity as occurs  at the analyte
wavelength measured. Background
correction is not required in cases of
line broadening where a background
correction measurement would
actually degrade the analytical result.
The possibility of additional
interferences named in 5.1 (and tests
for their presence as described in 5.2)
should also be  recognized and
appropriate corrections made.

3.   Definitions

3.1   Dissolved — Those elements
which will pass through a 0.45 /urn
membrane filter.

3.2   Suspended — Those elements
which are retained  by a  0.45 fjm
membrane filter.

3.3   Total — The concentration
determined on an unfiltered sample
following vigorous digestion (9.3), or
the sum of the dissolved plus
suspended concentrations. (9.1 plus
9.2.)

3.4   Total recoverable — The
concentration determined on an
unfiltered sample following treatment
with hot, dilute mineral acid (9.4).

3.5   Instrumental detection limit —
The concentration equivalent to a
signal, due to the analyte, which is
equal to three times the standard
deviation of a series of ten replicate
measurements of a  reagent blank
signal at the same wavelength.

3.6   Sensitivity —  The slope of the
analytical curve, i.e. functional
relationship between emission
intensity and concentration.

3.7   Instrument check standard — A
multielement standard of known
concentrations prepared by the
analyst to monitor ana verify
instrument performance on a daily
basis. (See 7.6.1)

3.8   Interference check sample —  A
solution containing  both interfering
and analyte elements of  known
concentration that can be used to
 verify background and interelement
 correction factors. (See 7.6.2)

 3.9  Quality control sample — A
 solution obtained from an outside
 source  having known, concentration
 values to be used to verify the
 calibration standards. (See 7.6.3)

 3.10  Calibration standards — a
 series of know standard solutions
 used by the analyst for calibration of
 the instrument (i.e., preparation of the
 analytical curve). (See 7.4)

 3.11  Linear dynamic range — The
 concentration  range over which the
 analytical curve  remains  linear.

 3.12  Reagent blank — A volume of
 deionized, distilled water containing
 the same acid matrix as the
 calibration standards carried through
 the entire analytical  scheme. (See
 7.5.2)

 3.13  Calibration blank — A volume
 of deionized, distilled water acidified
 with HN03 and HCI.  (See 7.5.1)

 3.14  Method of standard addition —
 The standard addition technique
 involves the use of the unknown and
 the unknown plus a  known amount of
 standard. (See 10.6.1)

 4.   Safety

 4.1  The toxicity or carcinogenicity of
 each reagent used in this method has
 not been precisely defined; however,
 each chemical compound should be
 treated  as a potential health  hazard.
 From this viewpoint, exposure  to
 these chemicals must be reduced to
 the lowest  possible  level by whatever
 means available. The laboratory is
 responsible for maintaining a current
 awareness file of OSHA regulations
 regarding  the safe handling of the
 chemicals specified in this method. A
 reference file of material data
 handling sheets should also  be made
 available to all personnel involved in
 the chemical analysis. Additional
 references to laboratory safety are
 available and have been identified
 (14.7, 14.8 and 14.9) for the
 information of the analyst.

5.   Interferences

5.1  Several types of interference
effects may contribute to inaccuracies
in the determination  of trace
elements. They can be summarized  as
follows:

5.1.1  Spectral interferences can be
categorized as  1) overlap of a spectral
line from another element; 2)
 unresolved overlap of molecular band
 spectra; 3) background contribution
 from continuous or recombination
 phenomena; and 4) background
 contribution from stray light from the
 line emission of high concentration
 elements. The first of these effects
 can be compensated by utilizing a
 computer correction of the raw data,
 requiring the  monitoring and
 measurement of the interfering
 element. The  second effect may
 require selection of an alternate
 wavelength. The third and fourth
 effects can usually be compensated by
 a background correction  adjacent to
 the analyte line. In addition, users of
 simultaneous multielement
 instrumentation must assume the
 responsibility of verifying the absence
 of spectral interference from an
 element that could occur in a sample
 but for which there is no channel in
 the instrument array. Listed in Table 2
 are some interference effects for the
 recommended wavelengths given in
 Table 1. The data in Table 2 are
 intended for use only as a
 rudimentary guide for the indication of
 potential spectral interferences.  For
 this purpose,  linear relations between
 concentration and intensity for the
 analytes and the interferents can be
 assumed.
  The interference information, which
 was collected at the Ames Laboratory,1
 is expressed at analyte concentration
 eqivalents (i.e. false analyte concen-
 trations) arising from 100 mg/L of the
 interferent  element. The suggested use
 of this information is as follows:
 Assume that arsenic (at 193.696  nm)
 is to be determined in a sample
 containing  approximately 10 mg/L of
 aluminum. According to Table 2, 100
 mg/L of aluminum would yield a  false
 signal for arsenic equivalent to
 approximately 1.3 mg/L.  Therefore,
 10  mg/L of aluminum would result in
 a false signal  for arsenic equivalent to
 approximately 0.13 mg/L. The reader
 is cautioned that other analytical
 systems may exhibit somewhat
 different levels of interference than
 those shown in Table 2, and that the
 interference effects must be evaluated
 for  each individual system.

  Only those  interferents  listed  were
 investigated  and the blank spaces in
 Table 2 indicate  that measurable inter-
 ferences were  not observed for the
 interferent concentrations listed in
 Table 3. Generally, interferences  were
 discernible if  they  produced  peaks or
 background  shifts corresponding to
 2-5%  of the  peaks  generated  by the
'Ames Laboratory, USDOE, Iowa Stale University,
 Ames Iowa 5001 1
                                      Dec. 1982
                                                                  Metals-2

-------
analyte concentrations also listed in
Table 3.
  At present, information on the listed
silver and potassium wavelengths are
not available but it has been reported
that second order energy from the
magnesium 383.231 nm wavelength
interferes with the listed potassium line
at 766.491 nm.

5.1.2  Physical interferences are
generally considered to be effects
associated with the sample nebuliza-
tion and transport processes. Such
properties as change in viscosity and
surface tension can cause significant
inaccuracies especially in samples
which  may contain high dissolved
solids and/or acid concentrations. The
use of a peristaltic pump may lessen
these interferences. If these types of
interferences are operative, they must
be reduced by dilution of the sample
and/or utilization of standard addition
techniques. Another problem which
can occur from high dissolved solids
is salt  buildup at the tip of the
nebulizer. This affects aersol flow-rate
causing instrumental drift. Wetting
the argon prior to nebulization, the
use of a tip washer, or sample dilution
have been used to control  this
problem. Also, it has been reported
that better control of the argon flow
rate improves instrument
performance. This is accomplished
with the use of mass flow controllers.

5.1.3  Chemical Interferences are
characterized by molecular compound
formation, ionization effects and
solute  vaporization effects. Normally
these effects are not pronounced with
the ICP technique, however, if
observed they can be minimized by
careful selection of operating
conditions (that is, incident power,
observation position, and so forth), by
buffering of the sample, by matrix
matching, and by standard addition
procedures. These types of
interferences can be highly dependent
on matrix  type and the specific
analyte element.

5.2 It is  recommended that
whenever a new or  unusual sample
matrix is encountered, a series of
tests be performed prior to reporting
concentration data for analyte
elements. These tests, as outlined in
5.2.1 through 5.2.4, will ensure the
analyst that neither positive nor
negative interference effects are
operative on any of the analyte el-
ements thereby distorting the
accuracy of the reported values.

5.2.1  Serial dilution—If the analyte
concentration is sufficiently high (min-
imally a factor of 10 above the instru-
mental detection limit after dilution),
an analysis of a dilution should agree
within 5 % of the original determina-
tion (or within some acceptable con-
trol limit (14.3) that has been estab-
lished for that matrix).  If  not, a
chemical or physical interference ef-
fect should be suspected.

5.2.2  Spike addition—The recovery
of a spike addition added at a
minimum level  of 10X the in-
strumental detection limit (maximum
100X) to the original determination
should be recovered to within 90 to
110 percent or  within the established
control limit for that matrix. If not, a
matrix effect should be suspected. The
use of a standard addition analysis
procedure can usually compensate for
this effect. Caution: The standard ad-
dition technique does not detect coin-
cident spectral overlap. If suspected,
use of computerized compensation,  an
alternate wavelength, or  comparison
with an alternate method is recom-
mended. (See 5.2.3)

5.2.3  Comparison with alternate
method of analysis—When investi-
gating a new sample matrix, compari-
son tests may be performed with other
analytical techniques such as atomic
absorption  spectrometry,  or other
approved methodology.

5.2.4   Wavelength scanning of
analyte line region—If the appropriate
equipment is available, wavelength
scanning can be performed to detect
potential spectral interferences.

6.   Apparatus

6.1  Inductively Coupled Plasma-
Atomic Emission Spectrometer.

6.1.1  Computer controlled atomic
emission spectrometer with background
correction.

6.1.2  Radiofrequency generator.

6.1.3  Argon gas supply, welding
grade or better.

6.2  Operating conditions — Because
of the differences between various
makes and models of satisfactory
instruments,  no detailed  operating
instructions can be provided. Instead,
the analyst should follow the
instructions provided by the
manufacturer of the particular
instrument. Sensitivity, instrumental
detection limit,  precision, linear dy-
namic range, and interference effects
must be investigated and established
for each individual analyte line on that
particular instrument. It is the
responsibility of the analyst to verify
that the instrument configuration and
operating conditions used satisfy the
analytical requirements and to
maintain quality control data
confirming instrument performance
and analytical results.

7.   Reagents and standards

7.1  Acids used in the preparation
of standards and for sample processing
must be ultra-high purity grade or
equivalent. Redistilled acids are
acceptable.

7.1.1  Acetic acid, cone, (sp gr 1.06).

7.7.2  Hydrochloric acid,  cone, (sp gr
1.19).

7.1.3  Hydrochloric acid,  (1 +1): Add
500 mL cone. HCI  (sp gr 1.19) to 400
ml_ deionized, distrilled water and
dilute to  1 liter.

7.1.4  Nitric acid, cone, (sp gr 1.41).

7.1.5  Nitric acid,(1 +1): Add 500 mL
cone.  HNO3 (sp.  gr 1.41) to 400 mL
deionized, distilled water and dilute to
1 liter.

7.2 Dion/zed, distilled water: Prepare
by passing distilled water through  a
mixed bed of cation and anion ex-
change resins. Use deionized, distilled
water for the preparation of all
reagents,  calibration standards and as
dilution water. The purity of this water
must be equivalent to ASTM Type  II
reagent water of Specification D 1193
(14.6).

7.3 Standard stock solutions may be
purchased or prepared from ultra high
purity grade chemicals or  metals. All
salts must be dried for 1 h at 105°C
unless otherwise specified.
(CAUTION: Many  metal salts are ex-
tremely toxic and may be fatal if swal-
lowed. Wash hands  thoroughly  after
handling.) Typical  stock solution  pre-
paration procedures follow:
7.3.1  Aluminum solution, stock,  1
mL  = 100^ Al: Dissolve 0.100 g of
aluminum metal  in an acid mixture of 4
mL  of (1 +1) HCI and 1 mL of cone. HNO3
in a beaker. Warm gently to effect
solution. When solution is complete,
transfer quantitatively to a liter flask,
add an additional 10 mL of (1 + 1) HCI
and dilute to  1,000 mL with deionized,
distilled water.

7.3.2  Antimony solution stock, 1  mL
= 100//g  Sb: Dissolve 0.2669 g K(SbO)
C4H406 in deionized distilled water,
add 10 mL (1+1) HCI and dilute
to 1000 mL with deionized, distilled
water.
                                       Metals-3
                                                                  Dec. 1982

-------
 7.3.3  Arsenic solution, stock,  1 mL =
 100 /jg As: Dissolve 0.1 320 g of As203
 in 100 mL of deiomzed, distilled water
 containing 0.4 g NaOH. Acidify the
 solution with 2 mL cone. HMOs and
 dilute to 1,000 mL with deionized,
 distilled water.

 7.3.4  Barium solution, stock, 1 mL
 = 100 /jg Ba: Dissolve 0.1516 g BaCI2
 (dried at 250°C for 2 hrs) in 10 mL
 deionized, distilled water with 1 mL
 (1+1) HCI. Add 10.0 mL(1+1) HCI
 and dilute to 1,000 mL with deionized,
 distilled water.

 7.3.5  Beryllium  solution, stock, 1
 mL = WO fjg Be: Do not dry. Dis-
 solve  1.966 g BeSO* • 4 4H2O, in
 deionized, distilled water, add 10.0 mL
 cone.  HNO3 and dilute to 1,000 mL
 with  deionized, distilled water.

 7.3.6  Boron solution, stock,  1  mL
 = 100 /ug B: Do not dry. Dissolve
 0.5716 g anhydrous H3BO3 in deionized
 distilled water  dilute to 1,000 mL.
 Use a reagent  meeting ACS specifica-
 tions, keep the bottle tightly stoppered
 and store in a desiccator to prevent
 the entrance of atmospheric moisture.

 7.3.7  Cadmium  solution, stock, 1
 rnL = 100 /ug Cd: Dissolve 0.1142 g
 CdO  in a minimum amount of (1 + 1)
 HN03.  Heat to increase rate of dis-
 solution. Add 10.0 mL cone. HNO3
 and dilute to 1,000 mL with deionized,
 distilled water.

 7.3.8  Calcium solution,  stock, 1  mL
 = 100 /ug Ca: Suspend 0.2498 g
 CaC03 dried at 180°C for 1 h  before
 weighing in deionized, distilled water
 and dissolve cautiously with a  min-
 imum amount  of (1 + 1) HIMO3. Add
 10.0 mL cone  HN03 and dilute to
 1,000 mL with deionized, distilled
 water.

 7.3.9  Chromium solution, stock, 1
 mL = 100 fjg Cr: Dissolve 0.1 923
 g of CrOs in deionized, distilled
 water.  When solution is complete,
 acidify with 10 mL cone. HNO3 and
 dilute to 1,000 mL with deionized,
 distilled water.

 7.3.10 Cobalt solution, stock,  1
 mL = 100yug Co: Dissolve 0.1000 g
 of cobalt metal in  a  minimum amount
 of (1+1) HN03. Add 10.0 mL (1 + 1) HCI
 and dilute to 1,000 mL with deionized,
 distilled water

7.3.1 1   Copper solution, stock,  1
mL = 100 fjg Cu: Dissolve 0.1252 g
CuO in a minimum amount of (1 + 1)
HN03. Add 10.0 mL cone.  HNO3 and
dilute to 1,000  mL with deionized,
distilled water.
 7.3.12  Iron solution, stock, 1  mL
 =  100 fjg Fe: Dissolve 0.1430 g
 Fe2O3 in a warm mixture of 20 mL
 (1+1) HCI and 2 mL of cone. HN03.
 Cool, add an additional  5 mL of cone.
 HNO3 and  dilute to 1000 mL with
 deionized,  distilled water.

 7.3.13  Lead solution,  stock, 1  mL
 =  100 fjg Pb: Dissolve 0.1599 g
 Pb(N03)? in minimum amount of
 (1+1) HN03. Add 10.0 mL cone. HN03
 and dilute  to 1,000 mL  with deionized,
 distilled water.

 7.3.14  Magnesium solution, stock, 1
 mL = 100 /jg Mg: Dissolve 0.1658 g
 MgO in a minimum amount of (1 + 1)
 HN03. Add 10.0 mL cone. HN03 and
 dilute to 1,000 mL with deionized,
 distilled water.

 7.3.15  Manganese solution, stock, 1
 mL = 100/jg Mn: Dissolve 0.1000 g
 of manganese metal in  the acid  mix-
 ture 10 mL cone HCI and 1  mL cone.
 HN03, and dilute to 1,000 mL with
 deionized,  distilled water.

 7.3.16  Molybdenum solution, stock,
 1  mL = 100/jg Mo:  Dissolve  0.2043 g
 (NH^MoCU in deionized, distilled
 water and  dilute to 1,000 mL.

 7.3.17  Nickel solution, stock,  1
 mL - 100 /jg Ni: Dissolve 0.1000 g
 of nickel metal in 10 mL hot cone.
 HN03, cool and  dilute to 1,000 mL
 with deionized, distilled water.

 7.3.18  Potassium solution, stock, 1
 mL = 100 fjg K:  Dissolve 0.1907 g
 KCI, dried at 110°C, in deionized,
 distilled water dilute to  1,000 mL.

 7.3.19  Selenium solution, stock, 1
 mL = 100 /jg Se: Do not dry.  Dissolve
 0.1727 g H2Se03 (actual assay 94.6%)
 in deionized, distilled water  and dilute
 to 1,000 mL.

 7.3.20  Silica solution, stock, 1 mL
 =  100 /jg Si02: Do not dry. Dissolve
 0.4730 g Na2SiO3 • 9H2O in deionized,
 distilled water. Add 10.0 mL cone.
 HN03 and  dilute to  1,000 mL with
 deionized,  distilled water.

 7.3.27  Silver solution, stock, 1
 mL - 100 /jg Ag: Dissolve 0.1 575 g
 AgN03 in 100 mL of deionized, dis-
 tilled water and 10 mL cone. HINOa.
 Dilute to 1,000  mL with deionized,
 distilled water.

 7.3.22  Sodium solution, stock, 1
 mL = 1 00 /jg Na: Dissolve 0.2542 g
 NaCI in  deionized, distilled water.
Add  10.0 mL cone. HN03 and dilute
to  1,000 mL with deionized,  distilled
water.
7.3.23  Thallium solution, stock, 1
mL = 100yug Tl:  Dissolve 0.1303 g
TIN03 in deionized, distilled water.
Add 10.0 mL cone. HNO3 and dilute
to 1,000 mL with deionized, distilled
water.

7.3.24  Vanadium solution,  stock, 1
mL = 100 fjg V: Dissolve 0.2297
NH4VO3 in  a minimum amount of
cone. HN03. Heat to increase rate
of dissolution. Add 10.0 mL cone.
HN03 and dilute to 1,000 mL with
deionized, distilled water.

7.3.25  Zinc solution, stock, 1 mL
= 100 fjg Zn: Dissolve 0.1245 g ZnO
in a minimum amount of dilute HN03.
Add 1 0.0 mL cone. HN03 and dilute
to 1,000 mL with deionized, distilled
water.

7.4  Mixed calibration standard so-
lutions—Prepare mixed calibration
standard solutions by combining ap-
propriate volumes of the stock solu-
tions in volumetric flasks. (See 7.4.1
thru 7.4.5) Add 2 mL of (1 + 1)
HCI and dilute to 100 mL with
deionized, distilled water. (See Notes
1 and 6.) Prior to preparing the mixed
standards,  each stock solution should
be analyzed separately to determine
possible spectral interference or the
presence of impurities. Care  should
be taken when preparing the mixed
standards that the elements  are com-
patible and stable. Transfer the mixed
standard solutions to a FEP fluoro-
carbon or unused polyethylene bottle
for storage. Fresh mixed standards
should be prepared as needed with
the realization that concentration can
change on  aging. Calibration stand-
ards  must be initially verified using
a quality control sample  and  moni-
tored weekly for stability (See 7.6.3).
Although not specifically required,
some typical calibration standard com-
binations follow when using  those
specific wavelengths listed in Table
1.

7.4.1  Mixed standard solution I—
Manganese, beryllium, cadmium, lead,
and zinc.

7.4.2  Mixed standard solution II—
Barium, copper,  iron, vanadium, and
cobalt.


7.4.3  Mixed standard solution III—
Molybdenum, silica, arsenic,  and
selenium.
7.4.4  Mixed standard solution IV—
Calcium, sodium, potassium, alumi-
num, chromium and nickel.
                                       Dec. 1982
                                                                    Metals-4

-------
7.4.5  Mixed standard solution V—
Antimony, boron,  magnesium, silver,
and thallium.
  NOTE 1: If the addition of silver
to the recommended acid combination
results in an initial precipitation,
add 1 5 mL of deionized distilled
water and warm the flask until the
solution clears. Cool and dilute to 100
mL with  deionized, distilled water. For
this acid combination the silver con-
centration should be limited to 2
mg/L.  Silver under  these conditions
is stable in a tap water matrix
for  30  days. Higher  concentrations
of silver  require additional HCI.

7.5  Two types of blanks are required
for the analysis. The calibration blank
(3.13) is  used in establishing the
analytical curve while the reagent
blank (3.12) is used to correct for
possible  contamination resulting from
varying amounts of  the acids used in
the sample processing.

7.5. /  The calibration blank is pre-
pared by diluting 2 mL of (1 + 1) HN03
and 10 mL of (1 + 1)  HCI to 100 mL
with deionized, distilled water. (See
Note 6.) Prepare a sufficient quantity
to be used to flush the system be-
tween  standards and samples.

7.5.2  The reagent blank must con-
contain all the reagents  and in the
same volumes as used in the pro-
cessing of the samples. The reagent
blank must be carried through the
complete procedure and contain the
same acid concentration in  the final
solution  as the sample solution
used for  analysis.

7.6  In addition to the calibration
standards, an instrument check stan-
dard (3.7), an interference check
sample (3.8) and a quality control
sample (3.9) are also required for the
analyses.

7.6.1  The instrument check standard
is prepared by the analyst by com-
bining  compatible elements at a con-
centration equivalent to  the midpoint
of their respective calibration curves.
(See 12.1.1)

7.6.2  The interference check sample
is prepared by the analyst in the
following manner. Select a
representative sample which contains
minimal  concentrations  of the
analytes of interest  by known con-
centration of interfering elements that
will provide an  adequate test of the
correction factors. Spike the sample
with the elements of interest at the
approximate concentration  of either
100 /ug/L or 5 times the estimated
detection limits given in Table 1. (For
effluent samples of expected high
concentrations, spike at an
appropriate level.) If the type of
samples analyzed are varied, a
synthetically prepared sample may be
used if the above criteria and intent
are met. A limited supply of a
synthetic interference check sample
will be available from the Quality
Assurance Branch of EMSL-
Cincinnati. (See 12.1.2)

7.6.3   The quality control sample
should be prepared in the same acid
matrix as the calibration standards
at a concentration near 1 mg/L and in
accordance with the instructions
provided by the supplier. The Quality
Assurance Branch of EMSL-Cincinnati
will either  supply a quality control
sample or  information where one of
equal quality can be procured. (See
12.1.3)

8.   Sample handling an
   preservation

8.1  For the determination of trace
elements, contamination and loss are
of prime concern. Dust in the labora-
tory environment, impurities in
reagents and  impurities on  laboratory
apparatus which the sample contacts
are all sources of potential
contamination. Sample containers can
introduce either positive or negative
errors in the measurement of trace
elements by (a) contributing con-
taminants through leaching or surface
desorption  and (b) by depleting
concentrations through adsorption.
Thus the collection and treatment of
the sample prior to analysis requires
particular attention. Laboratory
glassware including the sample bottle
(whether polyethylene, polyproplyene
or FEP-fluorocarbon) should be
thoroughly  washed with detergent
and tap water; rinsed with (1+1) nitric
acid, tap water, (1+1)  hydrochloric
acid, tap and finally deionized, distilled
water in that  order (See Notes 2  and
3).
NOTE 2:  Chromic acid may be useful to
remove organic deposits from glass-
ware; however, the analyst should be
be cautioned that the glassware must
be thoroughly rinsed with water to
remove the last traces of chromium.
This is especially important if chromium
is to  be included in the analytical
scheme. A  commercial product, NOCH-
ROMIX, available from Godax Labor-
atories,  6 Varick St., New York, NY
10013, may be used in place of
chromic acid. Gnomic acid should not
be used with  plastic bottles.
NOTE 3: If it can be documented through
an  active analytical quality control
program  using spiked samples and re-
agent blanks, that certain steps in the
cleaning  procedure are not required for
routine samples, those steps may be
eliminated from the procedure.

8.2  Before collection of the sample a
decision  must be  made as to the type
of data desired, that  is dissolved,
suspended or total, so that the appro-
priate preservation and pretreatment
steps may be accomplished. Filtration,
acid preservation,  etc., are to be per-
formed at the time the sample is
collected or as soon  as possible
thereafter.

8.2.1  For the determination of dis-
solved elements the  sample must be
filtered through a 0.45-/um membrane
filter as soon as practical after collec-
tion. (Glass or plastic filtering appara-
tus are recommended to avoid possi-
ble contamination.) Use the first 50-
100 mL to rinse the  filter flask. Dis-
card this portion and collect the
required volume of filtrate. Acidify the
filtrate with (1+1) HNO3 to a pH of 2
or less. Normally, 3 mL of (1+1) acid
per liter should be sufficient to pre-
serve the sample.

5.2.2  For the determination of sus-
pended elements  a measured volume
of unpreserved sample must be fil-
tered through a 0.45-jum membrane
filter as soon  as practical after
collection. The filter plus suspended
material should be transferred to a
suitable container for storage and/or
shipment. No preservative is required.

5.2.3  For the determination of total
or total recoverable elements, the
sample is acidified with (1 +1) HNOa
to pH 2 or less as soon as possible,
preferable at the time of collection.
The sample is not filtered before
processing.

9.   Sample Preparation
9.1  For the determinations of dis-
solved elements, the  filtered,
preserved sample  may often be
analyzed  as received. The acid matrix
and concentration of  the samples and
calibration standards must be the
same. (See Note 6.) If a precipitate
formed upon acidification of the
sample or during transit or storage, it
must be redissolved before the
analysis by adding additional acid
and/or by heat as described in 9.3.

9.2  For the determination of sus-
pended elements, transfer the mem-
brane filter containing the insoluble
material to a 150-mL Griffin beaker
and add 4 mL cone. HNOa. Cover the
                                      Met a Is-5
                                                                 Dec. 1982

-------
beaker with a watch glass and heat
gently. The warn acid will soon dis-
solve the membrane.
  Increase the temperature of the
hot plate and digest the material.
When the acid has nearly evaporated,
cool the beaker and watch glass and
add another 3 mL  of cone. HN03.
Cover and continue heating until the
digestion is complete, generally indi-
cated by a light colored digestate.
Evaporate to near dryness (2 mL), cool,
add 10 mL HCI (1+1) and 1 5 mL
deionized, distilled water per 100 mL
dilution and warm the beaker gently
for  15 min. to dissolve any precipi-
tated or residue material. Allow to
cool, wash down the watch glass and
beaker walls with deionized distilled
water and filter the sample to remove
insoluble material  that could clog the
nebulizer. (See Note  4.) Adjust the
volume based on the expected con-
centrations of elements  present. This
volume will vary depending on the
elements to be determined (See Note
6). The sample is now ready for
analysis. Concentrations so determined
shall  be reported as "suspended."
NOTE 4: In place of filtering, the
sample after diluting and mixing may
be centrifuged or allowed to settle  by
gravity overnight to remove insoluble
material.

9.3   For the determination of total
elements,  choose a measured, volume
of the well mixed acid preserved
sample appropriate for the expected
level of elements and transfer to a
Griffin beaker. (See Note 5.) Add 3  mL
of cone. HNOa. Place the beaker on
a hot plate and evaporate to near dry-
ness cautiously, making certain that
the sample does not  boil  and that no
area of the bottom of the beaker is
allowed to go dry. Cool the beaker and
add another 5 mL portion of cone.
HNOs. Cover the beaker with a watch
glass and return to the hot plate.
Increase the temperature of the hot
plate so that a gentle reflux action
occurs. Continue heating, adding addi-
tional acid as necessary,  until the
digestion is complete (generally indi-
cated when the digestate is light
in color or does not change in appear-
ance with continued  refluxing.) Again,
evaporate to near dryness and cool
the beaker. Add 10 mL of H1 HCI
and 1 5 mL of deionized, distilled
water per  100 mL  of final solution
and warm the beaker gently for 1 5
min. to dissolve any precipitate or
residue  resulting from evaporation.
Allow to cool, wash down the beaker
walls and watch glass with deionized
distilled water and filter  the sample to
remove  insoluble material that could
clog the nebulizer. (See Note 4.) Adjust
the sample to a predetermined volume
based on the expected concentrations
of elements present. The sample is
now ready for analysis (See Note 6).
Concentrations  so determined shall be
reported as "total."
NOTE 5: If low determinations of
boron are critical, quartz glassware
should be use.
NOTE 6: If the sample analysis solution
has a different acid  concentration
from that given  in 9.4, but does  not
introduce  a physical interference or
affect the analytical result, the same
calibration standards may be used.

9.4  For the determination of total
recoverable elements, choose a  mea-
sured volume of a well mixed, acid
preserved sample appropriate for the
expected level of elements and trans-
fer to a  Griffin beaker. (See Note 5.)
Add 2 mL of (1+1) HN03 and  10  mL
of (1 + 1) HCI to the sample and heat
on a steam bath or hot plate  until the
volume  has been reduced to  near 25
mL making certain the sample does
not boil. After this treatment, cool
the sample and  filter to remove  inso-
luble material that could clog the
nebulizer.  (See Note 4.) Adjust the
volume to  100 mL and mix. The sample
is now ready for analysis. Concentra-
tions so determined  shall be reported
as "total."

10.  Procedure

10.1  Set up instrument with proper
operating parameters established in
6.2. The instrument must be  allowed
to become thermally stable before be-
ginning. This usually requires at least
30 min. of operation prior to calibra-
tion.

10.2  Initiate appropriate operating
configuration of computer.

10.3  Profile and calibrate instru-
ment according  to instrument
manufacturer's  recommended
procedures, using the typical  mixed
calibration standard solutions
described in 7.4. Flush the system
with the calibration  blank (7.5.1)
between each standard. (See Note 7.)
(The use of the average intensity of
multiple exposures for both
standardization and  sample analysis
has been found  to reduce random
error.)
NOTE 7:  For boron concentrations
greater  than 500 /ug/L extended flush
times of 1  to 2 min.  may be required.

10.4  Before beginning the sample
run, reanalyze the highest mixed
calibration standard  as if it were a
sample. Concentration values obtained
should not deviate from the actual
values by more than ± 5 percent
(or the established control  limits
whichever is lower). If they do, follow
the recommendations of the  instru-
ment manufacturer to correct for this
condition.

10.5  Begin the sample run flushing
the system with the calibration blank
solution (7.5.1) between each sample.
(See Note 7.) Analyze the instrument
check standard (7.6.1) and the calibra-
tion  blank (7.5.1) each 10 samples.

10.6  If it has been found that
method of standard addition  are
required, the following procedure is
recommended.

10.6.1  The standard addition tech-
nique (14.2) involves preparing new
standards  in the sample matrix by
adding known  amounts of standard to
one  or more aliquots of the processed
sample solution. This technique com-
pensates for a sample constituent that
enhances or depresses the analyte
signal thus producing a different slope
from that of the calibration standards.
It will not correct for additive inter-
ference which causes a baseline shift.
The  simplest version of this technique
is the single-addition method. The
procedure is as follows. Two identical
aliquots of the sample solution, each
of volume Vx, are taken. To the
first (labeled A) is added a small
volume Vs of a standard analyte
solution of concentration cs.  To the
second (labeled B) is added the same
volume Vs of the solvent. The analy-
tical signals of A and B are measured
and  corrected  for nonanalyte signals.
The  unknown  sample concentration
C* is calculated:

           cx =   SBVsCS
               (SA-SB)Vx

where SA and  SB are the analytical
signals (corrected for the blank) of
solutions A and B, respectively. Vs
and  cs should  be chosen so that SA
is roughly twice SB on the average. It
is best if Vs is  made much less than
Vx, and thus cs is much greater than
cx, to avoid excess dilution of the
sample matrix. If a separation or
concentration  step is used, the
additions are best made first and
carried through the entire procedure.
For the results from this technique to
be valid, the following limitations
must be taken  into consideration:
1. The analytical curve must  be linear.
2. The chemical form of the analyte
added must respond the same as the
analyte in the  sample.
                                      Dec. 1982
                                                                  Metals-6

-------
3. The interference effect must be
constant over the working range of
concern.
4. The signal must  be corrected for
any additive interference.

11.   Calculation

11.1  Reagent blanks (7.5.2) should
be subtracted from  all samples. This is
particularly important for  digested
samples requiring large quantities of
acids to complete the digestion.

11.2  If dilutions were performed,
the appropriate factor must be applied
to sample values.

11.3  Data should be rounded to the
thousandth place and all results
should be reported  in mg/L up to
three  significant figures.

12.  Quality Control
(Instrumental)

12.1  Check the instrument
standardization by analyzing
appropriate quality  control check
standards as follow:

12.1.1  Analyze an appropriate
instrument check standard (7.6.1)
containing the elements of interest at
a frequency of 10%. This  check
standard is used to determine
instrument drift. If agreement is not
within ±5% of the expected values or
within the established control limits,
whichever is lower, the analysis is out
of control. The analysis should be
terminated, the problem corrected,
and the instrument recalibrated.
  Analyze the  calibration  blank (7.5.1)
at a frequency of 10%. The result
should be within the established
control limits of two standard devia-
tions of the mean value. If not, repeat
the analysis two more times and
average the three results. If the
average is not within the  control  limit,
terminate the analysis,  correct the
problem and recalibrate the
instrument.

12.1.2  To verify interelement and
background correction factors analyze
the interference check sample (7.6.2)
at the beginning, end, and at periodic
intervals throughout the sample run.
Results  should fall  within the
established control  limits of 1.5 times
the standard deviation of  the mean
value. If not, terminate  the analysis,
correct the problem and recalibrate
the instrument.

12.1.3  A quality control sample
(7.6.3) obtained from an outside
source must first be used for the
initial verification of the calibration
 standards. A fresh dilution of this
 sample shall be anlayzed every week
 thereafter to monitor their  stability. If
 the results are not within ±5% of the
 true value listed for the control
 sample,  prepare a new calibration
 standard and recalibrate the
 instrument. If this does not correct the
 problem, prepare a new stock
 standard and a new  calibration
 standard and repeat  the calibration.

 Precision and Accuracy

 13.1  In an EPA round robin phase 1
 study, seven laboratories applied the
 ICP technique to acid-distilled water
 matrices that had been dosed with
 various metal concentrates. Table 4
 lists the true value, the mean reported
 value and the mean  % relative
 standard deviation.

 References
 1.  Winge, R.K., V.J. Peterson,  and
 V.A. Fassel, "Inductively Coupled
 Plasma-Atomic Emission
 Spectroscopy: Prominent Lines," EPA-
 600/4-79-017.

 2.  Winefordner, J.D.,  "Trace
 Analysis: Spectroscopic Methods for
 Elements," Chemical Analysis, Vol.
 46, pp. 41-42.

 3.  Handbook for Analytical Quality
 Control in Water and Wastewater
 Laboratories, EPA-600/4-79-019.

 4.  Garbarino, J.R. and Taylor, H.E.,
 "An  Inductively-Coupled Plasma
 Atomic Emission Spectrometric
 Method for Routine Water Quality
 Testing," Applied Spectroscopy 33,
 No. 3(1979).

 5.  "Methods for Chemical Analysis of
 Water and Wastes," EPA-600/4-79-
 020.

 6. Annual Book of ASTM Standards,
 Part 31.

 7.  "Carcinogens - Working With
 Carcinogens," Department of Health,
 Education, and Welfare, Public Health
 Service, Center for Disease  Control,
 National  Institute for  Occupational
 Safety and Health,  Publication No. 77-
 206, Aug. 1977.

 8. "OSHA Safety and  Health Stan-
dards, General Industry," (29 CFR
 1910), Occupational Safety and Health
Administration, OSHA 2206, (Revised,
January 1976).

9. "Safety in Academic Chemistry
Laboratories,  American Chemical So-
ciety Publication, Committee on
Chemical Safety, 3rd  Edition, 1979.
                                      Met a Is-7
                                                                 Dec. 1982

-------
Table 1.    Recommended Wavelengths ' and Estimated Instrumental
Detection Limits
Element
Aluminum
Arsenic
Antimony
Barium
Beryllium
Boron
Cadmium
Calcium
Chromium
Cobalt
Copper
Iron
Lead
Magnesium
Manganese
Molybdenum
Nickel
Potassium
Selenium
Silica (SiOz)
Silver
Sodium
Thallium
Vanadium
Zinc
Wavelength, nm
308.215
193.696
206.833
455.403
313.042
249. 773
226.502
317.933
267.716
228.616
324. 754
259.940
220.353
279.079
257.610
202.030
231.604
766.491
196.026
288.158
328.068
588.995
190.864
292.402
213.856
Estimated detection
limit, ng/Lz
45
53
32
2
0.3
5
4
10
7
7
6
7
42
30
2
8
15
see3
75
58
7
29
40
8
2
  1 The wavelengths listed are recommended because of their sensitivity and
   overall acceptance. Other wavelengths may be substituted if they can
   provide the needed sensitivity and are treated with the same corrective
   techniques for spectral interference. (See 5.1.1.J.
  2The estimated instrumental detection limits as shown are taken from
   "Inductively Coupled Plasma-Atomic Emission Spectroscopy-Prominent
   Lines, "EPA-600/4-79-017. They are given as a guide for an instrumental!
   limit. The actual method detection limits are sample dependent and may yary
   as the sample matrix varies.
  ^Highly dependent on operating conditions and plasma position.
                                     Dec. 1982                    Metals-8

-------
Table 2.    Analyte Concentration Equivalents (mg/L) Arising From Interferents at the TOO mg/L Level

Analyte         Wavelength, nm                                    Interferent

Aluminum
Antimony
Arsenic

308.215
206.833
193.696
Al Ca
0.47 —
1.3 —
Cr
2.9
0.44
Cu
	
Fe
0.08
Mg
—
Mn Ni
0.21 —
Ti
.25
V
1.4
0.45
1.1
Barium
Beryllium
Boron

Cadmium
Calcium
Chromium

Cobalt
Copper
Iron

Lead
Magnesium
Manganese

Molybdenum
Nickel
Selenium

Silicon
Sodium
Thallium

Vanadium
Zinc
455.403
313.042
249.773

226.502
317.933
267.716

228.616
324.754
259.940

220.353
279.079
257.610

202.030
231.604
196.026

288.158
588.995
190.864

292.402
213.856
0.04    —
0.17    —
 —      0.02
0.005   —

0.05    —

0.23    —



0.30    —
                0.08


                0.03
0.11
0.01
0.07    —
               0.32     —

               0.03     —
               0.01    0.01
               0.003    —

               0.005    —
               0.003    —
                —     0.02
                0.04     —
                0.04     —

                —     0.03

                0.12     —
0.13    —      0.25     —
0.002   0.002   —      —

0.03    —      —      —

0.03    -      -      -
                                                              0.04    0.05
0.03    0.03
—      0.04

0.15    —
0.05    0.02
                                               0.07    0.12
                0.05    —     0.005   —
                —     0.14     —      —
                                —      0.01
                               0.08    —
                                                              0.02    —
                                       0.29    —
 Table 3.    Interferent and Analyte  Elemental Concen-
           trations Used for Interference Measurements
           in Table 2.
Analytes    (mg/L)
           Interferents
            (mg/L)
Al
As
B
Ba
Be
Ca
Cd
Co
Cr
Cu
Fe
Mg
Mn
Mo
Na
Ni
Pb
Sb
Se
Si
TI
V
Zn
10
10
10
1
1
1
10
1
1
1
1
1
1
10
10
10
10
10
10
1
10
1
10
Al
Ca
Cr
Cu
Fe
Mg
Mn
Ni
Ti
V













1000
1000
200
200
WOO
WOO
200
200
200
200













                                    Met a Is-9
                                                              Dec. 1982

-------
Table 4.



Element
Be
Mn
V
As
Cr
Cu
Fe
Al
Cd
Co
Ni
Pb
Zn
Se
ICP Precision and Accuracy Data
Sample # 7

True
Value
vg/L
750
350
750
200
150
250
600
700
50
500
250
250
200
40
Mean
Reported
• Value
V9/L
733
345
749
208
149
235
594
696
48
512
245
236
201
32

Mean
Percent
RSD
6.2
2.7
1.8
7.5
3.8
5.1
3.0
5.6
12
10
5.8
16
5.6
21.9

True
Value
M9/L
20
15
70
22
10
11
20
60
2.5
20
30
24
16
6
Sample #2
Mean
Reported
Value
H9/L
20
15
69
19
10
11
19
62
2.9
20
28
30
19
8.5


Mean
Percent
FISD
9.8
6.7
2.9
23
18
40
15
33
16
4.1
11
32
45
42


True
Value
V9/L
180
100
170
60
50
70
180
160
14
120
60
80
80
10
Sample #3
Mean
Reported
Value
W/L
176
99
169
63
50
67
178
161
13
108
55
80
82
8.5


Mean
Percent
RSD
5.2
3.3
1.1
17
3.3
7.9
6.0
13
16
21
14
14
9.4
8.3
Not all elements were analyzed by all laboratories.
                                       Dec. 1982
                                                                  Metals-JO

-------
                               United States
                               Environmental Protection
                               Agency
                               Environmental Monitoring and
                               Support Laboratory
                               Cincinnati OH 45268
                               Research and Development
£EPA
Test  Method
                               Chromium,  Dissolved
                               Hexavalent  (Atomic
                               Absorption, Furnace
                               Technique)—Method  218.5
                               1.   Scope and Application

                               1.1  This method covers the
                               determination of dissolved hexavalent
                               chromium (Cr6+) in drinking and
                               surface waters. The method may also
                               be applicable to certain domestic and
                               industrial wastes after filtration
                               provided that potential interferring
                               substances are  taken into account.
                               (See 4.1.)

                               1.2  The method may be used to
                               analyze samples containing from 5 to
                               100 /JQ of Cr6+ per liter. The range of
                               the method may be extended upward
                               by dilution.

                               2.   Summary of Method

                               2.1  The method is based on the
                               separation of Cr6+ from the sample by
                               coprecipitation  of lead chromate with
                               lead sulfate in a solution of acetic
                               acid. After separation, the supernate
                               is drawn off and the Cr5 precipitate
                               resolubilized in  nitric acid as trivalent
                               chromium (Cr3+) and quantified by
                               furnace atomic  absorption.

                               2.2  Hexavalent chromium may  also
                               be analyzed by  the chelation/
                               extraction technique (see Method
                               218.4 or the procedure described in
                               9.2 of the Atomic Absorption methods
                               found in this manual).
                               3.   Sample Handling and
                               Preservation

                               3.1  For sample handling, cleaning
                               glassware and the filtration procedure
                               see part 4.1 of the Atomic Absorption
                               Method section of this manual.

                               3.2  The sample must not be
                               preserved by acidification, but instead
                               transported and stored until time of
                               analysis at 4°C.

                               3.3  Stability of Cr6+ in environmental
                               samples is n ot completely understood
                               at this time. The chemical nature of
                               the sample matrix can have a definite
                               affect on the chemistry of chromium.
                               Therefore, the analysis should be
                               carried out as soon as possible but no
                               longer than 24 hours after collection.

                               4.   Interferences

                               4.1  The possible interference from
                               other elements which form stable
                               chromates is not known at this time.

                               4.2  Samples with either sulfate or
                               chloride concentrations above 1000
                               mg/liter should be diluted before
                               analysis.

                               4.3  The potential reduction of Cr6+
                               from highly reductive substances
                               increases as pH is lowered. When
                               sulfites and sulfides are present the
                               218.5-1
                                                     Dec. 1982

-------
sample aliquot taken for analysis
should be neutralized and aerated
before beginning.

5.   Instrument  Parameters
(General)

5.1   Drying Time and Temp: 30 sec-
125°C.

5.2   Ashing Time and Temp: 30 sec-
1000°C.

5.3   Atomizing Time and Temp: 10
sec-2700°C.

5.4   Purge Gas Atmosphere: Argon

5.5   Wavelength: 357.9nm

5.6   Other operating parameters
should be as specified by the
particular instrument manufacturer.
6.   Special Apparatus

6.1  Glassware

6.1.1  Filtering flask, heavy wall, 1
liter capacity

6.1.2  Centrifuge tubes, heavy duty,
conical, graduated, glass stoppered,
10 ml capacity

6.1.3  Pasteur pipets, borosilicate
glass, 5 % inches.

6.2  Centrifuge: any centrifuge
capable of reaching 2000 rpm and
accepting the centrifuge tubes
described in 6.1.2 may be used.

6.3  pH  Meter: a wide variety of
instruments are commercially
available and suitable for this work.

6.4  Test Tube Mixer: any mixer
capable of thorough vortex is
acceptable.
7.   Reagents

7.1  Lead Nitrate Solution: Dissolve
33.1 grams of lead nitrate, Pb(NO3)2
(analytical reagent grade), in deionized
distilled water and dilute to 100 ml_.

7.2  Ammonium Sulfate Solution:
Dissolve 2.7 grams of ammonium
sulfate, (NH4)2S04 (analytical reagent
grade), in deionized distilled water and
dilute to 100 ml.

7.3  Calcium Nitrate Solution:
Dissolve 1 1.8 grams of calcium
nitrate, Ca(NO3)2 • 4H2O (analytical
reagent grade), in deionized distilled
water and dilute to 100 mL.  1 ml =
20 mg Ca.
7.4  Nitric Acid, cone.: Distilled
reagent grade or equivalent to
spectrograde quality.

7.5  Acetic Acid, Glacial: ACS
reagent grade.

7.5.1  Acetic Acid, 10% (v/v):  Dilute
10 mL  glacial acetic acid to 100 mL
with deionized distilled water.

7.6  Ammonium Hydroxide, 10%
(v/v): Dilute 10 mL cone ammonium
hydroxide, NH4OH (analytical reagent
grade), to 100 mL with deionized
distilled water.

7.7  Hydrogen Peroxide, 30%:  ACS
reagent grade.

7.8  Potassium Dichromate Standard
Solution: Dissolve 2.8285 grams of
dried potassium dichromate, K2Cr2O7
(analytical reagent  grade), in deionized
distilled water  and  dilute to 1 liter. 1
mL = 1 mg  Cr (1000 mg/L)

7.9  Trivalent Chromium Working
Stock Solution:  To 50 mL of the
potassium dichromate standard
solution (78) add 1 mL of 30%  H202
(7.7) and 1  mL cone. HN03 (7.4) and
dilute to 100 mL with deionized
distilled water  1  mL = 0.5 mg Cr3+.
Prepare fresh monthly or as needed.
8.   Calibration

8.1  At the time of analysis prepare a
blank and a series of at least four
calibration standards from the Cr3+
working stock (7.9) that will
adequately bracket the sample. The
normal working range covers a
concentration range of 5 to  100 ug
Cr/L. Add to the blank and each
standard 1 mL 30% H202 (7.7), 5 mL
CONC HN03 (7.4), and 1 mL calcium
nitrate solution (7.3) for each 100 mL
of prepared solution before diluting  to
final volume. These calibration
standard should be prepared fresh
weekly or as needed.

8.2  The listed instrumental
conditions (5.)  and the stated
calibration concentration range are  for
a Perkin-Elmer HGA-2100 based on
the use of a 20/uL injection,  continous
flow purge gas and non-pyrolytic
graphite. The use of simultaneous
background correction is required for
both calibration and sample analysis.

9.   Procedure

9.1  Transfer  a 50 mL  portion of the
filtered sample to a  100mL Griffin
beaker and adjust to pH 3.5±0.3 by
adding 10% acetic acid  dropwise.
Record the volume of acid added and
adjust the final result to account for
the dilution.
Note: Care must be exercised not to
take the pH below  3. If the pH is
inadvertently  lowered to < 3, 10%
NH4OH (7.6) should be  used  to raise
the pH to above 3.

9.2  Pipet a  10  mL  aliquot  of  the
adjusted sample into a centrifuge tube
(6.1.2). Add 100/jL of the lead  nitrate
solution (7.1), stopper the tube, mix the
sample and allow to stand for 3min

9.3  After the formation of lead
chromate, retain the Cr3+ complex in
solution by addition of 0.5 mL glacial
acetic acid (7.5). Stopper and mix.

9.4  To provide adequate lead sulfate
for coprecipitation  add  100 mL
ammonium sulfate solution (7.2),
stopper and mix.

9.5  Place the stoppered centrifuge
tube in the centrifuge,  making sure
that the tube  is properly
counterbalanced. Start the centrifuge
and slowly increase the speed to
2000 rpm in small increments over a
period of 5 min. Centrifuge the sam-
ple at 2000 rpm for 10 min.
Note 2: The speed  of the centrifuge
must be increased slowly to  insure
complete coprecipitation.

9.6  After centrifuging remove  the
tube and draw off the  supernate using
the apparatus detailed  in Figure 1.  As
the pasteur pipet is lowered  into the
tube the supernate is sucked into the
filtering flask. With care the  supernate
can be withdrawn to within
approximately 0.1  mL above  the
precipitate.

9.7  To the remaining precipitate add
0.5 mL cone HNO3 (7.4), 100/uL  30%
H2O2 (7.7) and 100/A. calcium nitrate
solution (7.3). Stopper the tube and
mix using a vortex mixer to disrupt
the precipitate and solubilize the lead
chromate. Dilute to 10mL, mix and
analyze in the same manner as  the
calibration standard (8.2).

9.8  For the  general furnace
procedure and calculation, see
"Furnace  Procedure" part 9.3 of the
Atomic Absorption Methods  section of
this manual.

10.  Verification

10.1   For every sample matrix
analyzed verification is necessary to
determine that neither  a reducing
condition nor a chemical interference
affecting precipitation  is present. This
                                      Dec. 1982
                                                                   218.5-2

-------
must be accomplished by analyzing a
second 10ml_ aliquot of the pH
adjusted filtrate (9.1) spiked with Cr6*
(7.8). The amount of spike added
should double the concentration found
in the original aliquot. Under no
circumstance should the increase be
of less than 3Qug Cr6VL. To verify the
absence of an interference the spike
recovery should be between 85% and
115%.

10.2  If the addition of the spike
extends the concentration beyond the
range of the calibration curve, the
analysis solution should be diluted
with blank  solution and the calculated
results adjusted accordingly.

10.3  If the verification indicates a
suppressive interference, the sample
should be diluted and reanalyzed.


11.  Analytical  Notes

11.1  Nitrogen should not be used as
a purge gas because of possible CN
band  interference.

11.2  The use of pyrolytic graphite
should be avoided when possible.
Generally, pyrolytic graphite resulted
in a more limited analytical working
range and in some situations an
enhancement effect.

11.3  Pipet tips have been reported
to be  a possible source of
contamination. (See part 5.2.9 of the
Atomic Absorption Methods section of
this manual.)

11.4  The method of standard
addition should  not be required in as
much as the Cr6^ has been separated
from the original sample solution and
redissolved in a uniform matrix having
an absorption response coincident to
the calibration curve.

11.5  Data to be entered  into
STORET (No 01032) must be reported
as pg/L
12.3   Using Cincinnati, Ohio tap
water spiked at concentrations of
5,10, and 50 pg Cr6VL the standard
deviations were ± 0.7,  ±0.6, and
—0.6,  respectively Spike recovery at
all three levels was  102%.

12.4   A 1000 /jg Cr3VL standard
solution analyzed by this method
yielded a result of 8 /jg Cr6"/L with a
relative standard deviation of 19%.

12.5   The data from 5/jg Cr6VL tap
water spike was used to calculate
method detection limit (MDL) with
99% confidence as described in
"Trace Analyses for Wastewater," J.
Glaser, D.  Foerst, G. McKee, S.
Quave, W.  Budde, Environmental
Science and Technology, Vol.  15,
Number 12, page 1426, December
1981. The calculated MDL for
Cincinnati drinking water is 2.3 fjg/L.
12.  Precision and Accuracy

12.1  In a single laboratory (EMSL)
using a mixed industrial-domestic
waste effluent containing 22 pg
Cr6Vl_ and spiked with a
concentration of 50 /jg Cr6Vl_ the
standard  deviations were ± 1.0 and ±
2.7, respectively with a spike recovery
of 94%.

12.2  Recoveries of a 40 /jg Cr6VL
spike  in diluted  tannery and plating
waste effluents  were 96% and 93%,
respectively.
                                      218.5-3
                                                                 Dec. 1982

-------
                                 United States
                                 Environmental Protection
                                 Agency
                                 Environmental Monitoring and
                                 Support Laboratory
                                 Cincinnati OH 45268
                                 Research and Development
&EPA
Test  Method
                                 Sodium  (Atomic
                                 Absorption,  furnace
                                 technique)—Method  273.2
                                 Optimum Concentration Range: 1-30
                                 A
-------
  6.  Data to be entered into STORET
     must be reported as /ug/L.

Precision and Accuracy
  1.  Precision and accuracy data are
     not available at this time.
                                      Dec. 1982                    273.2-2

-------
                                  United States
                                  Environmental Protection
                                  Agency
                                   Environmental Monitoring and
                                   Support Laboratory
                                   Cincinnati OH 45268
                                  Research and Development
AEPA
Test Method
                                  Acidity (Titrimetric)—
                                  Method  305.2
                                  1.  Scope and Application

                                  1.1 This method is applicable to
                                  rain, surface and other waters of pH
                                  less than 8.3.

                                  1.2 This method is a measure of the
                                  concentration of strong and weak
                                  acids that react with hydroxyl ions.
                                  This includes the dissolved gases that
                                  are present.

                                  1.3 The range of this method
                                  depends on the volume of sample
                                  titrated and upon the precision that
                                  the increments of titrant can be
                                  measured. If only 10 mL of sample is
                                  available for analysis,  it is necessary
                                  to use a 50//L syringe for dispensing
                                  the titrant in order to achieve a
                                  precision of less than  10/ueq/L

                                  2.  Summary of Method

                                  2.1 Samples are titrated with 0.02 N
                                  carbonate free NaOH solution. The
                                  end point is determined with a pH
                                  meter. Results are reported as
                                  microequivalents (fieq) per  liter.

                                  3.  Sampling Handling and
                                  Storage

                                  3.1 The sample container must be
                                  filled completely, sealed and stored at
                                  4°C. Care must be taken to minimize
                                  exposure of the sample to the
                                  atmosphere. Open the sample
                                  container immediately before analysis.

                                  3.2 Analysis should  be performed as
                                  soon as possible after collection.

                                  4.  Comments

                                  4.1 Samples with  an initial pH
                                  between 4.3 and 8.3 are subject to
                                  error due to the loss or gain of
                                  dissolved gases during sampling,
                                  storage and analyses.

                                  5.   Apparatus

                                  5.1  pH meter and electrode(s), see
                                  Method  150.1 or 150.2.

                                  5.2  Micro buret or micro syrings.

                                  5.3  Teflon or glass magnetic stirring
                                  bar.

                                  5.4  Magnetic stirrer.

                                  5.5  Beakers or flasks.

                                  6.   Reagents

                                  6.1  Standard sodium hydroxide
                                  solution, 1 N: Dissolve 40g NaOH in
                                  250 mL  distilled water. Cool and dilute
                                  to 1 liter with COa free distilled water.
                                  Store in a polyolefin bottle and fitted
                                  with a soda lime tube or tight cap to
                                  protect from atmospheric C02.

                                  6.2  Standard sodium hydroxide
                                  titrant, 0.02 N: Dilute 20.0 mL of 1 N
                                  NaOH with  C02-free distilled water to
                                  1 liter. Store in rubber stoppered
                                  bottle. Protect from atmospheric C02
                                  by using a soda lime tube.
                                  Standardize against an 0.02 N
                                  potassium acid phthalate solution
                                  prepared by dissolving 4.085 g of
                                  anhydrous KHC8H404 in COa free
                                  distilled  water and diluted to 1:1.

                                  7.   Procedure

                                  7.1  Pipet an appropriate aliquot of
                                  sample into beaker of flask containing
                                  a small teflon on glass stirring bar.
                                  Use extreme care to minimize the
                                  sample surface disturbance.
                                  305.2-1
                                                           Dec. 1982

-------
7.2  Immerse pH electrode(s) into
sample and stir at a rate that does not
cause sample surface disturbance.

7.3  Titrate with 0.02 N NaOH  (6.2)
to pH 8.3. Titration should be made as
quickly as possible to prevent
absorption of atmospheric CC>2.
Record volume of titrant.

8.   Calculation

8.1   Acidity,/jeq/L =1!!Ex NBx 105
                   ml_s

            /ueq/L = microequivalents
                    per liter

            mLB = ml of NaOH titrant

              mLs = mL of sample

            NB = normality of titrant

9.   Precision and Accuracy

9.1   Precision and accuracy data are
not  available.

References
1.  Seymour, M.D., Schubert, S.A.,
    Clayton, J.W. and Fernando, Q.,
    Variation in the Acid Content of
    Rain Water in the Course of a
    Single Precipitation, Water, Air
    and Soil Pollution 10(2): 147-161,
    Aug. 1978.
2.  Peden,  M.E. and Skowron, Ionic
    Stability of Precipitation Samples;
    Atmospheric Environment, Vol.
    12, pp.  2343-2349. 1978.
3.  USGS,  Methods for Collection  and
    Analysis of Water Samples  for
    Dissolved Minerals and Gases, p.
    39, (1970).
4.  Annual Book of ASTM Standards,
    part 31, "Water," p.  107, D1067,
    (1978).
5.  Standard Methods for the
    Examination of Water and
    Wastewater,  14th Edition, p. 273,
    Method 402 (1975).
                                     Dec. 1982                     305.2-2

-------
                                United States
                                Environmental Protection
                                Agency
                                Environmental Monitoring and
                                Support Laboratory
                                Cincinnati OH 45268
                                Research and Development
V-/EPA
Test  Method
                                Organic  Carbon,  Total
                                (low level)
                                (UV  promoted,  persulfate
                                oxidation)—Method 41 5.2
                                1.  Scope and Application

                                1.1  This method covers the
                                determination of total organic carbon
                                in drinking water and other waters
                                subject to the limitations in 1.3 and
                                5.1.

                                1.2 This instrument is designed for
                                a two-step operation to distinguish
                                between  purgeable and nonpurgeable
                                organic carbon. These separate values
                                are not pertinent to this method.

                                1.3 This method is applicable only  to
                                the carbonaceous matter which is
                                either soluble or has a particle size of
                                0.2 mm or less.

                                1.4 The applicable  range is from
                                approximately 50 /ug/L to 10 mg/L.
                                Higher concentrations may be
                                determined by sample dilution.

                                2.  Summary of Method
                                  A sample is combined with  1 mL of
                                acidified  persulfate reagent and
                                placed in a sparger. The sample is
                                purged with helium which transfers
                                inorganic COa and purgeable organics
                                to a CC>2  scrubber. The COz is
                                removed  with at least 99.9%
                                efficiency with a 2.5-minute purge.
                                The purgeable organics proceed
                                through a reduction system where the
                                gas stream is joined by hydrogen and
                                passed over a nickel catalyst which
                                converts  the  purgeable organic carbon
                                to methane. The methane is
                                measured by a flame ionization
                                detector. The detector signal is
                                integrated and displayed as the
                                concentration of purgeable organic
                                carbon.
                                  The sample is then transferred to a
                                quartz ultraviolet reaction coil where
                                the nonpurgeable organics are
                                subjected to intense ultraviolet
                                illumination in the presence of the
                                acidified persulfate reagent. The
                                nonpurgeables are converted to CO2
                                and transferred to a second sparger
                                where a helium purge transfers the
                                CC>2 to the reduction system and into
                                the detector.  The signal is integrated,
                                added to the  purgeable organic carbon
                                value, and displayed as the
                                concentration of total organic carbon.

                                3.  Definitions

                                3.1. Total organic carbon measured
                                by this procedure is the sum of the
                                purgeable organic carbon and the
                                nonpurgeable organic carbon as
                                defined in 3.2 and 3.3.

                                3.2  Purgeable organic carbon is the
                                organic carbon matter that is
                                transferred to the gas phase when the
                                sample is purged with helium and
                                which passes through the CO2
                                scrubber. The definition is instrument-
                                condition dependent.

                                3.3  Nonpurgeable organic carbon is
                                defined as that which remains after
                                removal of the purgeable organic
                                carbon from the sample containing
                                acidified persulfate reagent and which
                                415.2-1
                                                       Dec. 1982

-------
 is converted to CO2 under the
 instrument conditions.

 3.4  The system blank is the value
 obtained in 8.2 for an irradiated,
 recirculated reagent distilled water
 sample.

 4.   Sample Handling and
•Preservation

 4.1   Sampling and storage of
 samples must be done in glass
 bottles. Caution: Do not leave any
 headspace in the sample bottle as
 this may contribute to loss of
 purgeable organics.

 4.2  Because of the possibility of
 oxidation or bacterial decomposition of
 some components of aqueous
 samples, the lapse of time  between
 collection of samples and start of
 analysis should be kept to a minimum.
 Also, samples should be kept cool
 (4°C) and protected from sunlight and
 atmospheric oxygen.

 4.3  When analysis cannot be
 performed within two hours from time
 of sampling, the sample should be
 acidified to pH 2 with H2S04.  Note:
 HCI should not be used because it is
 converted to chlorine during the
 analysis. This causes damage to the
 instrument.

 5.  Interferences

 5.1   If a sample is homogenized to
 reduce the size of the paniculate
 matter, the homogenizing may cause
 loss of purgeable organic carbon, thus
 yielding erroneously low results.

 6.  Apparatus

 6.1   Apparatus for blending or
 homogenizing samples: A household
 blender or similar device that will
 reduce particles in the sample to less
 than 0.2 mm.

 6.2   Apparatus for Total Organic
 Carbon: The essential components for
 the apparatus used in this method
 are: A sparge assembly, flow
 switching valves, a pyrolysis furnace,
 quartz ultraviolet reactor coil,  reducing
 column, flame ionization detector,
 electrometer and  integrator. This
 method is based on the Dohrmann
 Envirotech DC-54 Carbon Analyzer.
 Other instruments having similar
 performance characteristics may be
 used.

 6.3   Sampling Devices: Any
 apparatus that will reliably  transfer
 10 ml_ of sample to the sparger. A 50
 mL glass syringe is  recommended
when analyzing samples with easily
purgeable organics so as to minimize
losses.

7.   Reagents

7.1   Reagent Distilled Water:
Distilled water used in preparation of
standards and for dilution of samples
should be ultra-pure to reduce the
magnitude of the blank. Carbon
dioxide-free, double distilled water is
recommended. The water should be
distilled from permanganate or be
obtained from a system involving
distillation and carbon treatment. The
reagent distilled water value must be
compared to a system blank
determined on a recirculated distilled
water sample. The total organic
carbon value of the reagent distilled
water should be  less  than 60/ug/L.
Purgeable organic carbon values of
the reagent distilled water should  be
less than 4 /ug/L.

7.2   Potassium hydrogen phthalate,
stock solution, 500 mg carbon/liter:
Dissolve  1.063 g of potassium
hydrogen  phthalate (Primary Standard
Grade) in  reagent distilled water (7.1)
and dilute to  1 liter.

7.3   Potassium hydrogen phthalate (2
mg/L): Pipet 4 mL of  potassium
hydrogen  phthalate stock solution
(7.2) into a one liter volumetric flask
and dilute to the mark with  reagent
distilled water (7.1).

7.4   Potassium hydrogen phthalate (5
mg/L): Pipet 1 mL of  potassium
hydrogen  phthalate stock solution
(7.2) into a 100 mL volumetric flask
and dilute to the mark with  reagent
distilled water (7.1).

7.5  Potassium hydrogen phthalate
(10 mg/L): Pipet  2 mL of potassium
hydrogen phthalate stock solution
(7.2) into a 100 mL volumetric flask
and dilute to the  mark with  reagent
distilled water (7.1).

7.6  Acidified Persulfate Reagent.
Place 100 mL of  reagent distilled
water (7.1) in a container. Add  5 g of
potassium persulfate. Add 5 g (3 mL)
of concentrated (85%) phosphoric
acid.

7.7  Carbonate-bicarbonate, stock
solution,  1000 mg carbon/liter: Place
0.3500 g of sodium bicarbonate and
0.4418 g of sodium carbonate in a
100 mL volumetric flask. Dissolve with
reagent distilled water (7.1)  and dilute
to the mark.

7.8  Carbonate-bicarbonate, standard
solution 50 mg/L: Place 5 ml of the
carbonate-bicarbonate stock solution
in a 100 mL volumetric flask and
dilute to the mark with reagent
distilled water (7.1).

8.  Procedure

8.1  Allow at least 30 minutes
warm-up time. Leave instrument
console on continuously when in daily
use, except for the ultraviolet light
source, which should be turned off
when not in use for more than a few
hours.

8.2 Adjust all gas flows,
temperatures and cycle times to
manufacturer's specifications. Perform
the "System Cleanup and Calibration"
procedure  in the manufacturer's
specifications each day. Recirculate a
sample of irradiated distilled water
until two consecutive readings within
10% of each other are obtained.
Record the last value for the system
blank. This value is a function of the
total instrument operation and should
not vary significantly from previous
runs. Reasons for significant changes
in the value should be identified.

8.3 Check the effectiveness of the
CO2 scrubber by analyzing the
carbonate-bicarbonate standard
solution(7.8).  Add 1 mL of acidified
persulfate  reagent (7.6) to 50 mL of
the solution. Transfer 10 mL of the
solution-with-reagent to the first
sparger and start the analysis cycle.
No response, or a very minor reading,
should be obtained from this solution.

8.4  Add 1 mL of acidified persulfate
reagent (7.6) to 50 mL of reagent
distilled water (7.1) blank, standards
7.3, 7.4,  and 7.5 and the samples.

8.5  Calibrate the analyzer as
follows:

5.5. 7  Run the reagent distilled water
(7.1) and 5.0 mg/L standard  (7.4):
Transfer 10 mL of the solution-with-
reagent to the first sparger and start
analyzer cycle

Ignore the meter reading for  the first
cycle

Transfer a second 10 mL of the
solution-with-reagent to the first
sparger and start the analysis cycle

Record the meter reading (see 9.1) of
the final carbon value for each of the
reagent distilled water (7.1) and the
standard (7.4).

If the meter reading is more than 25%
above or below the calculated value  of
standard 7.4,  reanalyze the standard
                                      Dec. 1982
                                                                   415.2-2

-------
 and set the calibration within 25%
 (8.5.4), reanalyze the system blank,
 and then begin 8.5.1 again. If the
 meter reading (see 9.1) is within 25%
 of the calculated value, continue to
 next step. The calculated value is
 defined in 8.5.2.

 8.5.2  Calculate the factor for the
 deviation of the instrument reading
 (see 9.1) for the standard (7.4) from
 the calculated value by:

 standard reading -
 calculated value _ PACJQR
 calculated value

 where the  calculated value is that
 value obtained by using the weight of
 potassium  hydrogen phthalate and
 does not include the carbon
 contributed by the reagent distilled
 water (7.1) with which it has been
 diluted.

 8.5.3  Calculate the adjusted reading
 by:
 calculated  value + (ROW - (FACTOR X
 ROW)) = ADJUSTED READING.
 where  ROW = mean reagent distilled
 water (7.1) value.

 8.5.4  Push in CALIBRATE button
 after READY light comes on and
 adjust the SPAN control  to the
 ADJUSTED READING calculated in
 8.5.3.

 8.6  Analyze the standards 7.3 and 7.5
 in order to check the linearity of  the
 instrument at least once each day:

 Transfer 10 ml_ of the  solution-with-
 reagent to  the first sparger and start
 analyzer cycle

 Ignore the  meter reading for the  first
 cycle

 Transfer a  second 10 mL of the
 solution-with-reagent to  the first
 sparger and start the analyzer cycle

 Record the meter reading (see 9.1) of
 the final carbon value for each of the
 standards 7.3 and 7.5.
   The range of concentration used for
 calibrating the  instrument and
 checking the linearity of  the
 instrument should be ascertained
 from a knowledge of the range of
 concentrations expected  from the
 samples. Standards for lower ranges
 can be prepared by diluting standards
 7.2, 7.3, and 7.4.
  Transfer 10 mL of the solution-with-
reagent to the  first sparger and start
analyzer cycle

Ignore the meter reading for the first
cycle

Transfer a second 10 mL of the
solution-with-reagent to the first
sparger and start the  analyzer cycle

Record the meter reading (see  9.1) of
the final carbon value  for each of  the
samples.

9.   Calculations

9.1  The values are read off the final
digital readout in /jg/L. The system
blank reading obtained  in 8.2  must be
subtracted from all reagent distilled
water, standard and sample readings.

10.  Precision and Accuracy

10.1   In a single laboratory (MERL),
using raw river water, centrifuged
river water, drinking water, and the
effluent from a  carbon column which
had concentrations of 3.11, 3.10,
1.79, and 0.07  mg/L total organic
carbon respectively, the standard
deviations from ten replicates were
±0.13, ±0.03,  ±0.02, and ±0.02
mg/L, respectively.

10.2  In a single laboratory (MERL),
using potassium hydrogen phthalate
in  distilled water at concentrations  of
5.0 and  1.0 mg/L total organic carbon,
recoveries were 80% and 91%,
respectively.

Bibliography
1.  Proposed Standard Method for
Purgeable and Nonpurgeable Organic
Carbon  in Water (UV-promoted,
persulfate oxidation method). ASTM
Committee D-19, Task Group
19.06.02.03 (Chairman R. J. Joyce),
January 1978.

2.  Operating Instruction Dohrmann
Envirotech, 3420 Scott  Boulevard,
Santa Clara, California  95050.

3.  Takahashi, Y., "Ultra Low Level
TOC Analysis of Potable Waters."
Presented at Water Quality
Technology Conference, AWWA, Dec.
5-8, 1976.
 8.7  Analyze the samples. Transfer
 10 mL of sample with reagent to the
 first sparger and start the analysis
 cycle.
a US. GOVERNMENT PRINTING OFFICE 1963-659-095/0569
                                       415.2-3
                                                                  Dec. 1982

-------