EPA/600/R-93/073
                                                   March 1993
AN X-RAY FLUORESCENCE SURVEY OF LEAD

    CONTAMINATED RESIDENTIAL SOILS

          IN LEADVILLE, COLORADO:
                 A CASE STUDY
                         by

                C.A. Kuharic and W.H. Cole
    Lockheed Environmental Systems and Technologies Company
                   Las Vegas, Nevada

                         and

                A.K. Singh and D. Gonzales
              University of Nevada, Las Vegas
         Harry Reid Center for Environmental Research
                   Las Vegas, Nevada
                     Project Officer

                   Kenneth W. Brown
  Technology Support Center for Monitoring and Site Characterization
          Environmental Monitoring Systems Laboratory
                   Las Vegas, Nevada
  ENVIRONMENTAL MONITORING SYSTEMS LABORATORY
       OFFICE OF RESEARCH AND DEVELOPMENT
       U.S. ENVIRONMENTAL PROTECTION AGENCY
             LAS VEGAS, NEVADA 89193-3478
                                           Printed on Recycled Paper

-------
                             NOTICE
The information in this document has been funded wholly or in part by the United
States Environmental Protection Agency under Contract No. 68-CO-0049 to the
Lockheed Environmental Systems and Technologies Company. It has been subjected
to the Agency's peer and administrative review, and it has been approved for
publication as an EPA document. Mention of trade names or commercial products
does not constitute endorsement or recommendation for use.
                                ii

-------
                      TABLE OF CONTENTS

NOTICE	• • • • «•

TABLE OF CONTENTS	 ..•"*

LIST OF FIGURES	  v

LIST OF TABLES	 vi

ACKNOWLEDGEMENTS  	vii

FOREWORD	  1

INTRODUCTION	  2
     Background	  2
     History 	  2
     Site Description	•	  3
     Migration of Waste Materials	  3
     The Role of EMSL-Las Vegas	  4

PRINCIPLES OF X-RAY FLUORESCENCE  	  4
     THE X-MET 880	  5
     THE KEVEX 770	  5

INSTRUMENT CO-CALIBRATION   		:  5

SAMPLE HANDLING	  8
     FIELD SAMPLING	  8
     SAMPLE PREPARATION	  8
          The First 850 samples	  8
          The Rest of the Samples	  9

DATABASE MANAGEMENT	  9

QUALITY CONTROL	  9
     INTERNAL CONSISTENCY	  10
     QUALITY CONTROL CHARTS	  10
     CLP VERSUS X-METs			  11
     PRECISION, ACCURACY AND DETECTION LIMITS 	  13
     TOTAL VERSUS ESTIMATION ERROR	  16
                      »
FALSE POSITIVES/FALSE NEGATIVES 	  16

-------
POPULATION TESTING



CONCLUSIONS	



REFERENCES  ...	
17



20



21
                             IV

-------
                               LIST OF FIGURES
Figure 1. Calibration results for the Kevex and X-Mets 1, 2, and 3	   7
Figure 2. Correlation plots between Kevex and moist/dry X-Met.	   10
Figure 3. Quality control charts for Kevex and X-Met #1	 .   12
Figure 4. Correlation plot between CLP and combined X-Mets, and Kevex	   13
Figure 5. Frequency distribution of %RSD for moist field duplicates	   15
Figure 6. False negatives and false positives	   18
                                        V

-------
                                LIST OF TABLES
Table 1.  False positives and false negatives	   17
Table 2.  List of results of t-test and Kolmogorov-Smirnov (KS) test	   19
                                        vi

-------
                           ACKNOWLEDGEMENTS
      The authors are indebted to Mr. Ken Wangerud, the Remedial Project Manager,
personnel from the Roy F. Weston, Inc., Denver office, and to Dr. Rex Bryan and Dr. John
Drexler for their valuable technical suggestions and advice.
                                       Vll

-------

-------
                                   FOREWORD
      Continuing concern over the adverse impacts to human health due to exposure to
lead has prompted characterization efforts at numerous sites across the United States. One
of the primary  potential exposure routes is  through  ingestion and  inhalation of lead
contaminated soils.  This problem can be serious at old mining and smelting sites, especially
in the western United States.

      The California Gulch Superfund Site  in Leadville, Colorado  was added to the
National Priority List (NPL) in  1983.  It  is an historic mining and smelting site that is
currently the focus of extensive studies on soil lead contamination and bioavailability.
Studies  of  ground and surface water impacts are also  in progress. Personnel from the
Environmental Protection Agencies (EPA) Environmental Monitoring Systems Laboratory
at Las Vegas (EMSL-LV), the Denver office of Roy F. Weston, Inc., and Geostat Systems,
Inc.  (GSI)  used  field-portable X-ray  fluorescence (FPXRF) to determine  the spatial
distribution of lead concentrations in residential soils.

      This report details the FPXRF program  sample collection, preparation, and analysis
procedures, database management, and program quality assurance efforts at Leadville. The
program clearly demonstrates that small, field portable XRF instrumentation can produce
large quantities of acceptable quality data in a timely and cost-efficient manner when used
properly.

      When combined with the results of  blood lead level and bioavailability studies, this
data can help to develop a true assessment  of the risks posed by lead in the residential soils
of Leadville.

-------
                                 INTRODUCTION
Background

       During the summer of 1991, over 3700 soil samples were collected and analyzed for
lead content. These samples were collected from the edge of alley right-of-ways (alleyways)
because of access difficulties, and from individual yards.  The samples were analyzed using
X-ray fluorescence spectrometry. A laboratory-grade Kevex 770 spectrometer was used to
corroborate the analyses from three field-portable X-Met 880 spectrometers, which were
then used to rapidly generate data of known quality.

       Forty-five characterized Leadville residential soils were available at the onset of the
survey. Ten of these samples were used to co-calibrate the Kevex 770 and three different
X-Met 880 instruments, and the remaining 35 were used to confirm the calibrations.  The
first 850 samples were measured  field  moist after being brought to  the field laboratory.
They were then oven-dried and sieved, and approximately six grams of each subsampled into
polyethylene X-ray cells. These prepared  samples were analyzed on the Kevex and the X-
Met 880s.  Some of these samples were  randomly subsampled for Contract Laboratory
Program (CLP) analysis.  Data comparability was demonstrated by correlating X-Met and
Kevex results to each other and then to the CLP results.

       All remaining field samples were analyzed field moist on one of the three X-Met 880
instruments. One sample from  each preparation batch of 30 samples was selected for Kevex
analysis and samples for CLP analysis were randomly selected from the entire sample suite.
Results of the survey indicated that the  X-Met 880s produced good quality data.

       From the beginning, two major issues were of concern. First, was whether a single
matrix model for the X-Mets would be adequate to analyze all soils within the area to be
surveyed.  Comparison of the X-Met and CLP data for the same samples indicated this to
be the case.  Second, was whether the  alleyway samples and the nearby residential yard
samples were of a single population. The  non-parametric Kolmogorov-Smirnov procedure
and the paired-wise sample t-test indicated  no significant differences in the alleyway and yard
samples; thus inferences could be made in  unsampled yards from nearby alleyway easement
samples.

History

       Mining activity in the vicinity of Leadville was traced to the 1859 discovery of placer
gold in California Gulch.  The decline  of the placer  deposits led  to the search  for and
discovery of lode gold in 1868. This discovery failed to reverse the decline, but in 1874 a
heavy mineral that had been interfering  with the placer operations was identified as silver-
bearing lead carbonate. The ensuing silver boom was such that by 1880, there were enough
mines to keep a number of smelters in operation.

-------
       Eventually, the mining and processing turned from the oxidized, carbonate ores to the
 less desirable sulfide ores of the more mundane metals, lead and zinc.  Activity since has
 been cyclic, following trends in the base metals markets. Total metal production to 1966 was
 about 24,000,000 tons.  There is currently only one active major operation in  the area,
 ASARCO's Black Cloud Mine.  The last smelter ceased operations in 1960.

       The extensive underground mining activity left a legacy of drainage tunnels, perhaps
 most notably the Yak tunnel, and numerous mine waste rock and tailings piles dotting the
 landscape, almost entirely to the east and upslope of the town. The processing  activities,
 which were scattered around the town, left behind large amounts of slag.  Owing to the
 relative  inefficiency of the recovery  methods of the  time, much of the waste material
 contains high amounts of lead and zinc, along with some other metals.

       A preliminary EPA site evaluation took place in 1982 and 1983, and  the site was
 placed on the NPL in 1983. The initial concern was on the effluent from the Yak tunnel.
 A surge pond and treatment system have since been constructed to deal with this  aspect of
 the mining legacy. However, as is sometimes the case around old mining districts, material
 has been taken from abandoned dumps for use as fill, slag has been used as railroad ballast,
 and it has been crushed for use on streets in the winter. This use, taken in conjunction with
 the spread of smelter dust and wind transport of fine grained particles from various dumps,
 prompted concern about the effects of lead on people living in the town, especially the
 children.

 Site Description

       The town of Leadville lies about 100 miles southwest of Denver. It is situated on the
 western slope of the Mosquito Range, just  east and upslope of the Arkansas River,  at an
 elevation of approximately  10,200 feet.  The mining activity was predominantly on the
 eastern side of town, along the drainages of California, Evans, Stray Horse, Oregon, Malta
 and  Georgia Gulches.   The area lies within the Colorado  Mineral Belt  and  is  highly
 mineralized(1).

       The soils in the affected area are sandy loams to gravelly sandy loams, with varying
 amounts of admixed  slags and tailings. The Leadville sandy loam predominates, with lesser
 amounts of Pierian and Troutville gravelly sandy loams present(2).  Large piles of slag and
 tailings are also scattered around the site.

 Migration of Waste Materials

       Contaminant  transport occurred from natural processes, such  as wind and water
 dispersion, but in Leadville as previously stated, the  problem was exacerbated by human
impacts, ranging from smelter stack emissions to slags that were crushed and spread on icy
roads and used for railroad ballast.  Tailings and mine waste rock were thoroughly mixed
into local soils by over a century of mining and commercial activities where these materials

-------
 were often used for fill material.
       In addition, many homes are built among the waste rock/tailings piles, or immediately
 adjacent to them. The tailings piles in particular are used as recreational areas by many off-
 road motorcycle riders and are criss-crossed with trails. Children also have access to these
 areas as unsanctioned playgrounds.

 The Role of EMSLr-Las Vegas

       At the request of EPA Region 8, an intensive effort was mounted by the EMSL-LV
 Technology Support Center to gather a large amount of data on the surficial (0-4") spatial
 distribution of lead in the residential soils of Leadville during the 1991 field season.  Owing
 to the rapid and economical approach afforded by FPXRF equipment and methodology, the
 EMSL-LV FPXRF team became an integral part of the program design group, along with
 personnel from  Roy  F.  Weston, Inc., Denver, and  Geostat Systems, Inc., of Golden,
 Colorado.

       The study was designed to achieve three primary goals.  First, to demonstrate the
 ability of the  FPXRF instrumentation to  generate quantitative data of known quality.
 Second, to gather sufficient data to allow the use of geostatistics to determine the optimal
 sample spacing for further sampling (to be determined  by Geostat Systems, Inc.). Third, to
 define areas of contamination in specific concentration ranges of <500 mg/kg, 500-1500
 mg/kg, and >1500 mg/kg.

       To achieve these goals prior to field work, EMSL-LV participated in the writing of
 a Sampling and Analysis Plan and a Quality Assurance Project Plan as part of the Workplan
 for XRF(3).  Once the plans were  written and accepted,  the  EMSL-LV mobile XRF
 laboratory was moved to Leadville, and a temporary field laboratory was set up for sample
 preparation and analysis.  In addition, a database management  system was designed and
 implemented to handle the large amount of data generated  by the XRF program.

       Once on site, the next step was to run a feasibility study to demonstrate that the data
 generated by the FPXRF instruments would meet the data quality objectives of the project.
 The remainder of this report is dedicated strictly to XRF aspects of the program.

                    PRINCIPLES OF X-RAY FLUORESCENCE

       XRF spectrometry is based on the principle that photons produced from an X-ray
 tube or radioactive source bombard the sample to produce fluorescence. The incident
 photons impinge on the electron cloud of the atom.  Among other events,  this process
 creates vacancies in one or more of the inner shells.  The vacancies cause instability within
 the atom. As the outer electrons seek stability by filling  the vacancies in the inner shells, the
 atom emits energies as X-ray photons. The emitted energy (fluorescence) from a particular
shell is characteristic of the atom in which it was produced and is equal to the difference in
bonding energy between the outer shell electron and the vacant shell. Most  elements under

-------
 the photon bombardment fluoresce simultaneously to produce a spectrum of characteristic
 radiation.  It is this spectrum that the XRF detector senses and counts.

       There  are two types of XRF spectrometers,  energy dispersive and  wavelength
 dispersive.  The principal differences  are  in the method of detection of the fluorescent
 energies of the specimen and the method of quantifying the analytes of interest. For more
 detailed information on X-ray fluorescence, the reader is referred to Jenkins et_al.(4).

 THE X-MET 880

       The X-Met 880 used on the Leadville site is a field-portable, energy dispersive XRF
 spectrometer marketed by Outokumpu Electronics, Inc., Langhorne, PA. The unit is self-
 contained, battery powered, microprocessor based, and weighs 8.5 kg. The surface analysis
 probe is specifically designed for field use.  The X-Met 880 is hermetically sealed and can
 be decontaminated with soap and water.  The probe includes two radioisotope sources,
 Americium-241 and Curium-244, a proportional tube counter, and the associated electronics.
 The source is protected by a Nuclear Regulatory Commission approved safety shutter.  The
 electronic  unit has thirty-two  calibration memories called "model".  Each  model can be
 independently calibrated  for as many as six elements.  Using a multivariate regression
 procedure with the proper isotope  sources, the instrument can be calibrated to measure
 elements from silicon to uranium. Unknown sample intensities are then compared to the
 calibration curves to yield quantitative  concentrations.

 THE KEVEX 770

       The Kevex 770 is a laboratory-grade, energy dispersive XRF instrument marketed by
 Fisons Instruments, San Carlos, CA.  The instrument contains a 198 watt, Rh anode, liquid-
 cooled X-ray tube, and a cryogenically cooled, lithium drifted silicon solid state  detector.
 Optimal excitation conditions for analytes ranging from Na to U can be achieved using tube
 direct  excitation  with  filters or using secondary  targets for essentially monochromatic
 excitation.  The higher resolution of the Kevex (over five times that of the X-Met) allows
 for excellent qualitative scans for analytes that interfere with the analytes of interest.

       Calibration curves were chosen over Fundamental Parameters (FP)1 for quantitating
 the analytes of interest. It was decided in committee that calibration curves based on a suite
 of site specific calibration standards would be a more robust approach than using FP.

                        INSTRUMENT CO-CALIBRATION

       Forty-five  residential soil samples from  a  1990  sampling effort by  Walsh  and
Associates, Denver, Colorado, were available with corresponding EPA CLP analytical results.
   ^Fundamental Parameters  is a mathematical approach which resolves analyte interferences and
produces quantitative results for all analytes in the sample.

-------
Each sample was poured from its eight ounce jar into a pan, dried overnight at 100 °C, and
then sieved through a two mm sieve. The sample was then rolled 20 times corner-to-corner
on a piece of kraft paper down the long axis of the subsequent ellipsoid shape of the pile
of soil.  Approximately ten, half gram samples were extracted from the pile of soil and
placed in a 31 mm polyethylene X-ray cell which was sealed with 0.2 mil thick polypropylene
film. Ten of these samples which spanned the full concentration range were used to "co-
calibrate" the four XRF instruments (site specific soil standards).  The remaining samples
were used to verify the accuracy of the calibration procedures. There were four instruments
used in this study: one Kevex 770 and three X-Met 880s. Correlations between CLP results
and the results from the four instrument verifications are shown in Figure 1.

      Site specific soil standards were used to compensate for physical (particle size, bulk
density, heterogeneity)  and  chemical (spectral) matrix effects  that impact instrument
response when analyzing soils.  Calibration curves for one matrix usually give incorrect
results when used to analyze  samples  of a different matrix.

-------
LEADVILLE PROJECT: INSTRUMENT CO-CALIBRATION
n=35; R-squar«=0.89
CLPPbwt*
0.8

0.7
o.e
0.5
0.4
0.3
0.2
0.1

0.0

I I
	 --i 	 1™ 	
! |
	 f~ 	
•


1 «
	 ;_ 	
i
+ * ^ %*
j£ | +
* +?" * •
T : :
: •

	
	



.1- -t

+





	 —




i-






	










:
I
••• 	
*



	






	
	




















0.0 0.1 0.2 0.3 0.4 0.5 0.8 0.7 0.8
KEVBCPbWCK
CLP VS. KEVEX FQH LEAD
LEADVILLE PROJECT! INSTRUMENT CO -CALIBRATION
n=35; R— square=0.83
OPPbwM
0 S

• . 0.7
0.6
0.5
0 4
0.3

0.1

n rt
• •
	 i 	 i 	



	



+
*
•%•
!

	 f- 	

1 *

•^ *±*
+ ^

•





	 =f-' 	

4*-



	




+



>





i i





+ 	







:
:
"'"•' 	 t 	








	













ViJ i i l 	 1 	 1 	 1 	 j 	 1
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 <3










8
XMET1 Fbwtti
CLP VS. XMET rfl FOR LEAD
LEAD1
CLPPbwM
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0.0.
iflLLE PROJECT: INSTRUMENT CO-CALIBRATION
n=35; R— square=0.85





^
- 	 1 	




w
^Jl. 	

.

+ 4






+
............
	

4-

^


, 	 _

...


II

4



.
	
,,,






0.0 0.1 O.Z 0.3 0.4 0.5 0.6 0.7 0.8
XMEI2Pb*rt%
CLP VS. XMET « FOR LEAD
LEAD1
CLPPbwt*
0.8
0.7
0.6
0.5
0.4
0.3
0.1
VILLE PROJECT: INSTRUMENT CO -CALIBRATION
n=35; R-square=0.87




	 +..



	

+





	
]
	 t 	 -
: 4-
|
I

* 1
**" * ]
	 { 	 -







+



	
	







O.O 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8
XMEI2Pb«vt%
CLP VS. XMET #3 FOR LEAD
Figure 1.  Calibration results for the Kevex and X-Mets 1, 2, and 3.

-------
       At Leadville, at least four different matrices were expected to occur: a carbonate
tailings matrix, a sulfide tailings matrix, slag, and local soil. An assumption was made that
since human activity over the last century thoroughly mixed these matrices in the residential
portion of the town, a single set of calibration curves could be used.  The ten calibration
samples came from locations scattered across the townsite, and the success of the calibration
confirmation and subsequent CLP corroboratory analyses indicated that the assumption was
valid.

                              SAMPLE HANDLING

       It was originally intended that once X-Met correlation to the Kevex was clearly shown
to be acceptable (i.e., R2 > 0.7) on a significant number of samples (800-1000), the X-Mets
would  be taken into the field for in situ analysis.  This was to be done for the sake of
expedience, but the field sampling crews proved so efficient that sample analysis continued
in the laboratory for the duration of the program.    Access to  residential   lots  was
extremely limited at the beginning of the program, so transects were surveyed on  public
properties which were mostly along the unpaved shoulders and edges of alley rights-of-ways,
usually only a few feet from backyard property lines.

FIELD SAMPLING

       Intrusive samples were  collected at 25-foot intervals  along the transects.  At each
sample location, a volume of soil approximately six inches in diameter and four inches deep
was thoroughly disaggregated with a pick and shovel. The soil was turned onto itself seven
times to reduce heterogeneity and to allow a surface measurement technique to represent
a volume.  This was the surface upon which  in situ analysis was originally intended to be
performed.  Approximately 300 to 500 grams of soil was scooped into a large plastic bag,
labeled and custody sealed, then double bagged and placed in a cooler for transport to the
field laboratory.     All samples were obtained under chain of custody. Upon arrival at the
field laboratory, all samples were  logged in and remained  under locked custody.

SAMPLE PREPARATION

The First 850 samples

      Each bag was opened and the field moist  samples were analyzed in the bag three
times (by placing the probe in the bag), thoroughly shaking the sample in the bag between
analyses.  This mode of analysis was deemed analogous to in situ analysis, if the  program
had gone to in situ analysis.

      Each sample was then transferred to a Pyrex loaf dish and oven dried overnight at
100 °C. The dried sample was passed through a  ten mesh (2 mm) sieve onto a  three by
three foot piece of kraft paper where the sample  was rolled  onto itself 20 times to reduce
heterogeneity.  Approximately six grams taken in 10-12 subsamples was placed into  a 31 mm

-------
 diameter polyethylene X-ray cell and sealed with 0.2 mil polypropylene film.  The cupped
 sample was then analyzed on the Kevex and again on one of the three X-Mets.  The
 remaining dried soil was put into archive, still under chain-of-custody. All Pyrex loaf dishes
 were decontaminated with an Alconox/distilled water solution, rinsed with distilled water, and
 air dried.

       All X-Met measurements were done in triplicate and the values averaged partly due
 to the relatively short acquisition time (30 seconds livetime), but mostly to obtain a more
 representative composite value of each sample.  A sample is measured only once on the
 Kevex due to the long acquisition time (200 seconds livetime2).

 The Rest of the Samples

       Once good correlation was established between X-Met and Kevex analysis (see Figure
 2), only one in ten samples was  dried, cupped, and analyzed on the Kevex.  The X-Met
 procedure  continued in the same  manner throughout  the rest  of the program; that is,
 samples were brought in by the field crews, analyzed in triplicate in the plastic bag, and then
 archived.

                           DATABASE MANAGEMENT

       A database management system was developed using the Statistical Analysis Software
 supported by the SAS Institute Inc., Gary, NC.  Screens were designed to look exactly like
 the data recording forms. Along with all the analytical data (all triplicate measurements and
 associated QC samples), other data including easting, northing, analytical  date and time,
 sample type, instrument ID, and  a unique serial number for each sample were tracked by
 the management system.

       Data were  entered  in  batches (30  samples  per batch)  and a$ each batch was
 completed, it was printed out and 'hand' checked against the data recording forms for entry
 errors. All changes made to the database were tracked by a comparison procedure, and a
 printout of each  set of changes made to the data set was put into the hard copy dataset for.
 documentation purposes.

                              QUALITY CONTROL

       As the database grew, comparison plots were  periodically generated between the
Kevex  and the X-Mets  to  assess the degree  of correlation, as was  done with the co-
calibration, thus demonstrating internal consistency within the data collection effort. Once
the CLP  corroboratory data were available, the same type of correlation was generated,
confirming the accuracy of the data.
   2At an average deadtime of 40 percent, this equates to 280 seconds acquisition time per sample.

                                        9

-------
INTERNAL CONSISTENCY

      Figure 2a shows the correlation between dry Kevex results and moist X-Met results
and 2b shows the correlation between dry Kevex and dry X-Met results. Note that when the
samples are dried and cupped and reanalyzed on the X-Met, the correlation improves only
five percent. The X and Y axis have been cut off at 1.0 weight  percent  for illustrative
purposes (21 points in excess of 1.0 weight percent are removed from 2a and the R2 drops
from 0.90 to 0.85; 17 points are removed from 2b with no change in R2).  These plots show
excellent correlations, and provide assurance of good internal consistency throughout the
program.
   LEADVttlE PROJECT DRY KEVEX VS MOIST XMET RESULTS
            n = 1314; R—square value = .85

 reVEXPbwflC
       1,0
       0.0
         O.O  0.1  0.2  0.3  0.4  0.5  0.6  0.7 0.8 0.9 1.0
                LEWvriuwhwtK
   LEADVILLE PROJECT: DRY KEVEX VS DRY XMET RESULTS
           n - 1100; R-aquaro value - J»
                                                 0.0!
                                                  O.O 0.1 0.2 0.3  0.4  0.5  0.6  0.7  0.8  0.9  1.0
                                                          LEAD valUM ki wt%
 (a)
(b)
           Figure 2. Correlation plots between Kevex and moist/dry X-Met.
QUALITY CONTROL CHARTS

       In order to monitor instrument stability, quality control check samples (QCCS) were
measured before and after each block of ten samples on each X-Met, and with each batch
of 13 samples on the Kevex3.  A low- and a mid-calibration range of QCCS selected were
used to verify the calibration curves.
       Kevex 770 has a 16 place autosampler.  Each batch of samples consisted of 1 tube flux monitor
standard, 2 QC standards, and 13 routine samples.
                                         10

-------
The low-range sample was used to calculate detection limits (discussed in section entitled
"PRECISION, ACCURACY AND DETECTION LIMITS" below), and the mid-range was
plotted in a control chart to monitor instrument stability.  These control charts are shown
in Figure 3.

       There are notable differences in the charts that need explanation.  There were
considerably more samples run on  X-Met  Number One than on the Kevex, thus the
difference in a  horizontal sense.  There was only one set of QCCS run  on the Kevex;
however, there were at least three sets run on X-Met Number One due to breakage of the
polypropylene film and loss of sample. What was not foreseen, mainly because the authors
had not engaged in an XRF survey of this magnitude, was that time would wreak havoc with
the QCCS.  The 0.2 mil polypropylene film, through which the samples are X-rayed,
developed small holes with time and the soil leaked out. Consequently, the film had to be
replaced periodically and, being highly electrostatic, it extracted portions of the very fine
fraction from the soil, thus biasing the standard.  Occasionally, the film would simply break
and another QCCS would have to be used. The bias between the different QCCS samples
could be seen as several distinct concentration groupings, the  first break being at run
number 117 (for X-Met Number One).  All of these potential sources of error manifested
themselves, particularly in the control chart for X-Met Number One, because the majority
of the samples were analyzed on this instrument. Also, the QCCS were analyzed in triplicate
on the X-Mets and  were shaken  and  tapped between each measurement.  The least
variability is evident from the Kevex chart because considerably fewer samples were run on
this instrument and samples were run only once, thus much less abuse of the polypropylene
film occurred.

       In retrospect, the QCCS should have been pulverized and pressed to pellets,
alleviating the biases with time. The accuracy of the X-Met is  clearly displayed in the
correlation between CLP and X-Met data (see Figure 4).  Ninety-six of the 140 samples sent
to the CLP were run on X-Met Number One (R2 = 0.86).

CLP VERSUS X-METs

      Figure 4 shows the correlation plots between combined X-Met and CLP and between
the Kevex and CLP.  These plots show very good agreement between XRF  measurements
and those of the CLP, thus supporting the high degree  of accuracy achieved in this field
survey. The higher R2 value for the less sophisticated X-Mets is probably  a result of the
triplicate measurement average accounting for the heterogeneity of the sample better than
the single, longer measurement used on the Kevex.
                                       11

-------
    LEADVULE PROJECT  INSTRUMENT CONTROL CHAHT
KEVBCPbXt*
      0.160
      0.140


      0.120


      0.100


      o.oso


      0.060


      0.040


      0.020


      0.000
               —I—
                60
—l	1	1	1	

 12O   180   2*0   3OO


 CONSECUTIVE HUN NUMBER
                                           360   420
                  LEAD veluBi for KEVEX
      Broken linos ot Oio mean and +/- 2 standard deviations
                                                            LEAEMLLE PROJECT: INSTRUMENT CONTROL CHART
                                    XUETI PbvrtK
                                          0.160
                                                             0.1 40 :
                                          0.100



                                          O.OBO



                                          0.060



                                          0.040



                                          0.020



                                          0.000
 0    60   120   180    240   300   360

           CONSECUTIVE RUN NUMBER



        LEAD values for XMETJ1
Broken lines trt mean and +/— 2 standard deviations
                                       420
LEADVU
0.160.
0.140
0.120.
0.100
0,080
0.060
0.040
0,020
0.000
Br
LE PROJECT: INSTRUMENT CONTROL CHART

hffljn,jjrj
rwpM


i 	 1 	 r 	 1 	 r— 	 i i
3 60 120 180 240 300 360 420
CONSECUTIVE RUN NUMBER
LEAD vtJuM for XUET #2
oken lines at mean and •+/- 2 standard deviations
LEACML
0.160-
0.140-
0.120.
0.100.
0.080.
0.060.
0.040-
0.020
0.000
Br
US PROJECT: INSTRUMENT CONTROL CHART
— j-i — ITPTI
__(i . I . .
ir~~
T^IT


3 60 120 180 240 300 360 420
CONSECUTIVE RUN NUMBER
LEAD vriuM ferXMET #3
oken lines ai mean and +/— 2 standard deviations
                  Figure 3.  Quality control charts for Kevex and X-Met #1.
                                                     12

-------
LE
cu>R>wt%
1.4
1.2
1.0
o.a
0.6
0.4
0.2
0.0!
ADMLLE PROJECT: CLP VS ALL 3 XMET RESULTS
n=140; R-square=0.88




•M
-(..-^M
jfft.



4-
++
: -f •+**
*•>«• +
^*'-t



+
4.4.^. . .
p+4f


i-

+
+.
...*...





•I-




+
h
+
+
+












0-0 0-2 0.4 0.6 0.8 1.0 1.2 1.4
XMETFbwt^
LEAD valuae h wHt
LJ
CLPPbwt*
1.4
1.2
1.0
0.8
0.6
0.4
0.2
EADVILUE PROJECT: CLP VS KEVEX RESULTS
n=140; R-squ
-------
       Using the American Chemical Society (ACS)(5) definition of detection limits4, the low-
calibration level of QCCS estimated a quantitation limit (QL) of approximately 0.16 wt%
lead.  Given the errors introduced by the unstable QCCS standards, this detection limit is
probably artificially high.  This assumption is supported by Figure 2a. At the origin of both
these plots, a "take off' point for X-Met detection can be seen on the Y-axis at or below 0.1
wt% lead.  Note that the QL goes down slightly for the dried samples. Based on Figure 2a,
an empirical QL for lead measurements in this program is.set at approximately 0.1 wt%
(1000 ppm).
   *The ACS defines a minimum instrument detection limit as 3 times the standard deviation of a series
of nonconsccutive blanks or low level check samples, and a quantitation limit as 10 times the standard
deviation.

                                          14

-------
FREQUENCY
i
220 +
i
!
J
200 +
i
i
i
180 •*•
i
i
i
i
i
160 +
i
i
i
i
140 -!•
i
120 +
i
i
i
i
i
100 +


1
1
1
80 +
|
1
J
i
60 +
i
i
i
40 +
i
j
i
20 +
1
1
!


****
****
****
****
****
****
****
****
****
****
****
****
****

****
****

****

****
****
****
****
****
****
****
****
****

****
****
****

****
****
****

****
****
****
****
****
****
****
****
****
****
****
****
****
****
2.5



















****
****
****
****
****
****
****

****
****
***'*

****
****
****

**** ****
**** ****
**** ****
**** ****
**** ****
**** ****
**** **** ****
**** **** ****
**** **** ****
**** **** ****
**** **** ****
**** **** ****
**** **** ****
**** **** ****
7.5 12.5 17.5




Moments

N 531
Mean 8.8
Variance 80.8
Std Dev 9.0
.Skewness 2.1
Kurtosis 5.5
Coef Var T02.0







Quantiles(Def=5)
100% Max 58.3 99% 43.8
75% Q3 11.9 95% 26.8
50% Med 5.9 90% 19.5
25% 01 2.6 10% 1.0
0% Min 0 5% 0.5
1% 0.1




















****
****
**** ****
**** **** **** **** ****
22.5 27.5 32.5 37.5 42.5 47.5 52.5 57.5
RSD_PB MIDPOINT
Figure 5.  Frequency distribution of %RSD for moist Held duplicates.
                                         15

-------
TOTAL VERSUS ESTIMATION ERROR

      The above precision and QL estimates seem high when compared to those obtained
from a  laboratory, but there were in excess of 3700 samples obtained in this program.
Decisions on the spatial distribution of lead in Leadville will not be based on the individual
samples, but rather on a geostatistical model of those samples.  Consider the following
equation:               .                                           .
                          
-------
Action Level (ppm)
500
1000
1500
2000
2500
3000
False Positive
2 (8.7%)
9 (22.0%)
13 (22.4%)
21 (26.6%)
13 (14.6%)
8 (8.4%)
False Negative
3 (2.5%)
2 (2.0%)
3 (3.6%)
6 (9.5%)
4 (7.5%)
8 (17.0%)
 Table l.   False positives and  false negatives generated  at  six
 action levels.
                        POPULATION TESTING

  ^       explained above,  residential access was extremely limited
 at  the beginning of the field season,  so the first 50%  or more of
 the samples came from shoulders of alleys in direct proximity with
 residences.    The obvious  question  was  whether the contaminant
 concentrations along the  alleyways  were significantly  different
 trom contaminant concentrations in nearby residential soils (i e
 are alley samples representative of yard material?).   To  answer
 this question, 38  'cells'  were  located where samples were  taken
 both within  alley rights-of-way and  in the nearby residential
 yards.  The paired-sample t-test and  the nonparametric Kolmogorov-
 Smirnov  procedures of the SAS  software were used to show  that
 with very  few  exceptions,  there is  no  significant difference
 between contaminant concentrations in the alleys and in the nearby
  .
     The results of the tests on the individual cells  are shown in
Table 2.  The test results are shown for test  size 0.05.

     TW<  testing   methods  were   employed,   one  a  parametric
            lPPl°a??  (2-sainple t Test) , the other a nonparametric
            test  (Kolmogorov-Smirnov Test).   The two approaches
were tried  because the  distribution  of data  values  was tending
toward the lognormal, but not strongly so.   Thus in order to cover
all  the bases,  both methods  were tried.   Of the  38  pairwise
comparisons  of  yard  and nearby   alleyway samples,  five  were
significantly different using the parametric test, while only two
were significantly different using the  nonparametric " test .   This
confirms the more nearly lognormal distribution of  values.   Both
                          c°nclusi°"' that «»  *» Populations are
                                17

-------


£
                      S
§
3
S
                                                                               (fl
                                                                              H
                                                                               0)

                                                                               s-
                                                                               g
                                                                              •H
                                                                              ts
                                                                               ti
                                                                               0)


                                                                              "
                                                                              I

                                                                              4J
                                                                               (tf
 0)


?

 S
 0)
 tp

 01


                                                   s
                                     !
                           e     £
                           i

                                    1 i
                                    IS

                      §
 §
 3
 8
                                                                              •H

                                                                              4J
                                                                              •H

                                                                               (0

                                                                               O

                                                                               04


                                                                               0)
                                                                               10
                                                                               w
                                                                               I
                                                                               0)
                                                                               c

                                                                               0)
                                                                               (fl
 vo

 0)


 I
 •H
                                     18

-------
Cell
Number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
t
Test
NSD
NSD
NSD
SD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
SD
NSD
NSD
NSD
NSD
NSD
KS
Test
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
Cell
Number
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38


t
Test
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
SD
NSD
NSD
NSD
NSD
NSD
SD
SD


KS
Test
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
NSD
SD
NSD
NSD
NSD
NSD
NSD
SD
NSD


Table 2.  List of results of t-test and Kolmogorov-Smirnov (KS) test. "NSD" means no
      significant difference, "SD" means significant difference.
                                        19

-------
                           CONCLUSIONS
The lead concentration data generated by FPXRF on residential soils in Leadville is
optimal for producing concentration isopleth maps that depict gross contamination
patterns across the townsite. The large number of sample points analyzed minimizes
errors in estimating values at unsampled points, yielding a more representative
depiction of contaminant distribution than a smaller data set of higher data quality.

A very large number of samples (>3700) were analyzed in a period of about three
months, allowing for cost-  and time-effective determination of spatial patterns  of
contamination distribution.

Precision and accuracy of  the data  generated are good, being ±27% and ±6%
respectively.  The levels of  false positives and negatives are also quite low.

The initial assumption that residential soils in Leadville were of a very similar matrix
(with respect to XRF) appears to have proven valid.  The close agreement between
XRF results and randomly chosen samples analyzed by CLP methods indicates that
the single matrix model used with the  X-Met 880  instruments was a reasonable
approach to dealing with a soil whose initial components were distinctly different, but
presumably well mixed by residential and mining activities over a long period of time-

Lead data from alleyway samples can be used to infer lead concentrations in the soil
of nearby yards. Statistical  analyses showed no significant differences between lead
values in backyards and values in nearby alleyways.

As a recommendation, any long-term future work (greater than two weeks) using
FPXRF should use pelletized QCCS.  It is virtually impossible to prepare identical
QCCS from loose soil, which is .falsely interpreted as inter-instrument bias.  With
extended use, the fine soil fraction leaks through  holes in the polypropylene biasing
the sample, which falsely translates  to  instrumental drift (intra-instrument bias).
Pulverization of the QCCS material prior to splitting minimizes inter-instrument bias,
and pressing the pulverized soil into a pellet produces a stable, homogeneous check
standard, thus removing the intra-instrument bias.
                                  20

-------
                                 REFERENCES
1.
2.
3.
4.
5.
6.
7.
Tweto, Ogden. 1968. Leadville District, Colorado, in Ore Deposits in the United
States. 1933/1967. J.D. Ridge, ed. American Institute of Mining, Metallurgical, and
Petroleum Engineers. New York, New York. pp. 681-705.

United States Department of Agriculture, Soil Conservation Service, in cooperation
with the Colorado Agricultural Experiment Station.  1975. Soil Survey of Chaffee-
Lake area, Colorado and Parts of Chaffee and Lake Counties.  United States
Government Printing Office, Washington, D.C.

Workplan  Soil Sampling and X-ray Fluorescence  Analysis, Volume  1, Leadville
Colorado. September 1991.  Roy F. Weston, Inc. Denver, Colorado.

Jenkins, Ron, R.W. Gould, and D. Gedcke.  Quantitative X-ray Spectrometry. 1981.
Marcel Dekker, Inc.  New York and Basel.  586 pages.

American Chemical Society.  1983.  Principles of Environmental Analysis. Analytical
Chemistry, Vol. 55. pp 2210-2218.

Flatman, G.T., E. Englund,  A. Yfantis.  1988.   Geostatistical Approaches to the
Design of Sampling  Regimes, in  L.H. Keith,  ed.   Principles of Environmental
Sampling. American Chemical Society, pp. 73-84.

United States Environmental Protection  Agency.  March  1987.  Data Quality
Objectives for Remedial Response Activities - Development Process.  EPA 540/G-
87/003. Washington, D.C.
                                       21
                                                 •&U.S. GOVERNMENT PRINTING OFFICE: 1993 - 750-002/80242

-------

-------